WorldWideScience

Sample records for radiation hard scintillator

  1. Development of radiation hard scintillators

    International Nuclear Information System (INIS)

    Markley, F.; Woods, D.; Pla-Dalmau, A.; Foster, G.; Blackburn, R.

    1992-05-01

    Substantial improvements have been made in the radiation hardness of plastic scintillators. Cylinders of scintillating materials 2.2 cm in diameter and 1 cm thick have been exposed to 10 Mrads of gamma rays at a dose rate of 1 Mrad/h in a nitrogen atmosphere. One of the formulations tested showed an immediate decrease in pulse height of only 4% and has remained stable for 12 days while annealing in air. By comparison a commercial PVT scintillator showed an immediate decrease of 58% and after 43 days of annealing in air it improved to a 14% loss. The formulated sample consisted of 70 parts by weight of Dow polystyrene, 30 pbw of pentaphenyltrimethyltrisiloxane (Dow Corning DC 705 oil), 2 pbw of p-terphenyl, 0.2 pbw of tetraphenylbutadiene, and 0.5 pbw of UVASIL299LM from Ferro

  2. Testing the radiation hardness of lead tungstate scintillating crystals

    CERN Document Server

    Shao, M; Li Chuan; Chen, H; Xu, Z Z; Wang, Z M

    2000-01-01

    Large Hadron Collider operation will produce a high radiation background. PbWO/sub 4/ crystals are selected as scintillators for the CMS electromagnetic calorimeter. To reach the precise requirement for energy measurements, a strict requirement for the radiation hardness is needed. In this paper, we present a method for evaluating the radiation hardness and its measurement. Results for several full size (23 cm length) lead tungstate crystals under Co/sup 60/ gamma - ray irradiation are given, investigating the light yield loss and its longitudinal uniformity. (8 refs).

  3. Development and Studies of Novel Microfabricated Radiation Hard Scintillation Detectors With High Spatial Resolution

    CERN Document Server

    Mapelli, A; Haguenauer, M; Jiguet, S; Renaud, P; Vico Triviño, N

    2011-01-01

    A new type of scintillation detector is being developed with standard microfabrication techniques. It consists of a dense array of scintillating waveguides obtained by coupling microfluidic channels filled with a liquid scintillator to photodetectors. Easy manipulation of liquid scintillators inside microfluidic devices allow their flushing, renewal, and exchange making the active medium intrinsically radiation hard. Prototype detectors have been fabricated by photostructuration of a radiation hard epoxy resin (SU-8) deposited on silicon wafers and coupled to a multi-anode photomultiplier tube (MAPMT) to read-out the scintillation light. They have been characterized by exciting the liquid scintillator in the 200 micrometers thick microchannels with electrons from a 90Sr yielding approximately 1 photoelectron per impinging Minimum Ionizing Particle (MIP). These promising results demonstrate the concept of microfluidic scintillating detection and are very encouraging for future developments.

  4. Radiation Hard and High Light Yield Scintillator Search for CMS Phase II Upgrade

    CERN Document Server

    Tiras, Emrah

    2015-01-01

    The CMS detector at the LHC requires a major upgrade to cope with the higher instantaneous luminosity and the elevated radiation levels. The active media of the forward backing hadron calorimeters is projected to be radiation-hard, high light yield scintillation materials or similar alternatives. In this context, we have studied various radiation-hard scintillating materials such as Polyethylene Terephthalate (PET), Polyethylene Naphthalate (PEN), High Efficiency Mirror (HEM) and quartz plates with various coatings. The quartz plates are pure Cerenkov radiators and their radiation hardness has been confirmed. In order to increase the light output, we considered organic and inorganic coating materials such as p-Terphenyl (pTp), Anthracene and Gallium-doped Zinc Oxide (ZnO Ga) that are applied as thin layers on the surface of the quartz plates. Here, we present the results of the related test beam activities, laboratory measurements and recent developments.

  5. The chemistry of two-component fluoride crystalline optical media for heavy, fast, radiation hard scintillators

    International Nuclear Information System (INIS)

    Sobolev, B.P.; Krivandina, E.A.; Fedorov, P.P.; Vasilchenko, V.G.

    1994-01-01

    Prospects for preparation of two-component dense optical materials for scintillators are shown, using data on phase diagrams of about 300 MF m - RF n (m, n ≤ 4) type systems, formed by metal fluorides. Primary characteristics (decay time and light output of luminescence, radiation hardness, etc.) of some multicomponent crystals are reported

  6. Radiation damage studies on polystyrene-based scintillators

    International Nuclear Information System (INIS)

    Britvich, G.I.; Peresypkin, A.I.; Rykalin, V.I.

    1991-01-01

    The radiation resistance of polystyrene-based scintillators containing various scintillation dopes is reported. All samples were irradiated to 137 Cs gamma rays in air at room temperature. The examination of radiation resistance of about thirty fluorescence compounds has been made. The most radiation-hard fluores are X25, X31, 3HF and M3HF. 1 fig.; 6 tabs

  7. Study of Radiation Hardness of Gd2SiO5 scintillator for Heavy Ion Beam

    CERN Document Server

    Kawade, K; Itow, Y; Masuda, K; Murakami, T; Sako,T; Suzuki, K; Suzuki, T; Taki, K

    2011-01-01

    Gd2SiO5 (GSO) scintillator has very excellent radiation resistance, a fast decay time and a large light yield. Because of these features, GSO scintillator is a suitable material for high radiation environment experiments such as those encountered at high energy accelerators. The radiation hardness of GSO has been measured with Carbon ion beams at the Heavy Ion Medical Accelerator in Chiba (HIMAC). During two nights of irradiation the GSO received a total radiation dose of 7 × 10$^5$ Gy and no decrease of light yield was observed. On the other hand an increase of light yield by 25% was observed. The increase is proportional to the total dose, increasing at a rate of 0.025%/Gy and saturating at around 1 kGy. Recovery to the initial light yield was also observed during the day between two nights of radiation exposure. The recovery was observed to have a slow exponential time constant of approximately 1.5 × 10$^4$ seconds together with a faster component. In case of the LHCf experiment, a very forward region ex...

  8. Radiation hardness of undoped BGO crystals

    International Nuclear Information System (INIS)

    Sahu, S.K.; Peng, K.C.; Huang, H.C.; Wang, C.H.; Chang, Y.H.; Hou, W.S.; Ueno, K.; Chou, F.I.; Wei, Y.Y.

    1997-01-01

    We measured the radiation hardness of undoped BGO crystals from two different manufacturers. Such crystals are proposed to be used in a small-angle calorimeter of the BELLE detector of the KEK B-factory. Transparency and scintillation light output of the crystals were monitored to see the effect of radiation damage. The crystals show considerable radiation hardness up to 10.2 Mrad equivalent dose, which is much higher than the maximum expected dosage of 500 krad per year of running at BELLE. (orig.)

  9. Scintillation properties and X-ray irradiation hardness of Ce3+-doped Gd2O3-based scintillation glass

    International Nuclear Information System (INIS)

    Liu, Liwan; Shao, Chongyun; Zhang, Yu; Liao, Xili; Yang, Qiuhong; Hu, Lili; Chen, Danping

    2016-01-01

    Ce 3+ -doped Gd 2 O 3 -based scintillation glasses are prepared within an air or CO atmosphere. The effects of fluorine, lutetium, barium, and the melting atmosphere on the optical properties, scintillation properties and irradiation hardness are studied. Absorption spectra, luminescence spectra under UV and X-ray excitation, and the X-ray radiation-induced spectra are presented. The results show that the density can be increased by doping with fluorine, lutetium and barium. The luminescence intensity decreases after X-ray irradiation. Because of charge transfer quenching, fluorine and lutetium enhance the UV-excited and X-ray excited luminescence intensity, but barium decreases. Moreover, fluorine and lutetium are advantageous to irradiation hardness while barium is not. In addition, a non-reducing atmosphere provides a higher irradiation hardness than a reducing atmosphere. Fluorine-doped glass is promising to enhance luminescence intensity, promote irradiation hardness, and increase the density.

  10. Influence of variable tungsten valency on optical transmittance and radiation hardness of lead tungstate (PWO) scintillation crystals

    CERN Document Server

    Burachas, S; Makov, I; Saveliev, Yu; Ippolitov, M S; Man'ko, V; Nikulin, S P; Nyanin, A; Vasilev, A; Apanasenko, A; Tamulaitis, G

    2003-01-01

    A new approach to interpret the radiation hardness of PbWO//4 (PWO) scintillators is developed by revealing importance of the inclusions of tungsten oxides WO//3//-//x with variable valency. It is demonstrated that the influence of the ionizing radiation on PWO is, in many aspects, similar to the effect of the high-temperature annealing in oxygenless ambient. In both cases, a valency change of the tungsten oxides is initiated and results in induced absorption and, consequently, in crystal coloration. In the PWO crystals doped with L//2O//3 (L = Y, La, Gd), the radiation hardness and the optical properties are mainly affected by inclusions of W//1//-//yL//yO//3//- //x (0 less than x less than 0.3) instead of inclusions of WO//3//- //x prevailing in the undoped samples. It is demonstrated that the radiation-induced bleaching and the photochromic effect of PWO are caused by phase transitions in the inclusions of tungsten oxide. Thermodynamic conditions for the phase transitions are discussed and the optimal oxid...

  11. Radiation hardness of plastic scintillators for the Tile Calorimeter of the ATLAS detector

    CERN Document Server

    Jivan, Harshna; The ATLAS collaboration

    2014-01-01

    The Tile Calorimeter of the ATLAS detector, is a hadronic calorimeter responsible for detecting hadrons as well as accommodating for the missing transverse energy that result from the p-p collisions within the LHC. Plastic scintillators form an integral component of this calorimeter due to their ability to undergo prompt fluorescence when exposed to ionising particles. The scintillators employed are specifically chosen for their properties of high optical transmission and fast rise and decay time which enables efficient data capture since fast signal pulses can be generated. The main draw-back of plastic scintillators however is their susceptibility to radiation damage. The damage caused by radiation exposure reduces the scintillation light yield and introduces an error into the time-of flight data acquired. During Run 1 of the LHC data taking period, plastic scintillators employed within the GAP region between the Tile Calorimeter’s central and extended barrels sustained a significant amount of damage. Wit...

  12. Studies for the LHCb SciFi Tracker - Development of Modules from Scintillating Fibres and Tests of their Radiation Hardness

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00341158

    The LHCb detector will see a major upgrade in the LHC long shutdown 2, which is planned for 2019/20. Among others, the tracking stations, currently realised as silicon strip and drift tube detectors, will be replaced by the Scintillating Fibre (SciFi) Tracker. The SciFi Tracker is based on scintillating fibres with a diameter of $\\text 250 \\mu m$, read out by multichannel silicon photomultipliers. The two major challenges related to the fibres are the radiation damage of the light guidance and the production of precise multi-layer fibre mats. This thesis presents radiation hardness studies performed with protons at the tandem accelerator at Forschungszentrum Garching and in situ in the LHCb cavern. The obtained results are combined with additional data of the LHCb SciFi group and two different wavelength dependent models of the radiation induced attenuation are determined. These are used to simulate the relative light yield, for both models it drops to $83 \\%$ on average at the end of the nominal lifetime of ...

  13. Thin-film-based scintillators for hard x-ray microimaging detectors: the ScinTAX Project

    Science.gov (United States)

    Rack, A.; Cecilia, A.; Douissard, P.-A.; Dupré, K.; Wesemann, V.; Baumbach, T.; Couchaud, M.; Rochet, X.; Riesemeier, H.; Radtke, M.; Martin, T.

    2014-09-01

    The project ScinTAX developed novel thin scintillating films for the application in high performance X-ray imaging and subsequent introduced new X-ray detectors to the market. To achieve this aim lutetium orthosilicate (LSO) scintillators doped with different activators were grown successfully by liquid phase epitaxy. The high density of LSO (7.4 g/cm3), the effective atomic number (65.2) and the high light yield make this scintillator highly applicable for indirect X-ray detection in which the ionizing radiation is converted into visible light and then registered by a digital detector. A modular indirect detection system has been developed to fully exploit the potential of this thin film scintillator for radiographic and tomographic imaging. The system is compatible for high-resolution imaging with moderate dose as well as adaptable to intense high-dose applications where radiation hard microimaging detectors are required. This proceedings article shall review the achieved performances and technical details on this high-resolution detector system which is now available. A selected example application demonstrates the great potential of the optimized detector system for hard X-ray microimaging, i.e. either to improve image contrast due to the availability of efficient thin crystal films or to reduce the dose to the sample.

  14. Radiation hardness of new Kuraray double cladded optical fibers

    International Nuclear Information System (INIS)

    Bedeschi, F.; Menzione, A.; Budagov, Yu.; Chirikov-Zorin, I.; Solov'ev, A.; Turchanovich, L.; Vasil'chenko, V.

    1996-01-01

    The radiation hardness of the new plastic scintillating and clear fibers irradiated by 137 Cs γ-flux and by pulsed reactor fast neutrons were investigated. All the studied fibers were of S-type (with S=70) and had a double cladding. Optical fibers degradation study after irradiation shows that the level of radiation hardness lower that what is expected from results of previous studies. 9 refs., 6 figs

  15. Studies for the LHCb SciFi tracker. Development of modules from scintillating fibres and tests of their radiation hardness

    Energy Technology Data Exchange (ETDEWEB)

    Ekelhof, Robert Jan

    2016-05-18

    The LHCb detector will see a major upgrade in the LHC long shutdown 2, which is planned for 2019/20. Among others, the tracking stations, currently realised as silicon strip and drift tube detectors, will be replaced by the Scintillating Fibre (SciFi) Tracker. The SciFi Tracker is based on scintillating fibres with a diameter of 250 μm, read out by multichannel silicon photomultipliers. The two major challenges related to the fibres are the radiation damage of the light guidance and the production of precise multi-layer fibre mats. This thesis presents radiation hardness studies performed with protons at the tandem accelerator at Forschungszentrum Garching and in situ in the LHCb cavern. The obtained results are combined with additional data of the LHCb SciFi group and two different wavelength dependent models of the radiation induced attenuation are determined. These are used to simulate the relative light yield, for both models it drops to 83% on average at the end of the nominal lifetime of the SciFi Tracker. A machine and techniques to produce multi-layer fibre mats were developed and optimised. Procedures for the production and alignment are described. These are implemented in the serial production of the SciFi modules which will start in the second quarter 2016.

  16. Scintillation counter based radiation dosimeter

    International Nuclear Information System (INIS)

    Shin, Jeong Hyun

    2009-02-01

    The average human exposure per year is about 240mrem which is come from Radon and human body and terrestrial and cosmic radiation and man-made source. Specially radiation exposure through air from environmental radiation sources is 80mrem/yr(= 0.01mR/hr) which come from Terrestrial and cosmic radiation. Radiation dose is defined as energy deposit/mass. There are two major methods to detect radiation. First method is the energy integration using Air equivalent material like GM counter wall material. Second method is the spectrum to dose conversion method using NaI(Tl), HPGe. These two methods are using generally to detect radiation. But these methods are expensive. So we need new radiation detection method. The research purpose is the development of economical environmental radiation dosimeter. This system consists of Plastic/Inorganic scintillator and Si photo-diode based detector and counting based circuitry. So count rate(cps) can be convert to air exposure rate(R/hr). There are three major advantages in this system. First advantages is no high voltage power supply like GM counter. Second advantage is simple electronics. Simple electronics system can be achieved by Air-equivalent scintillation detector with Al filter for the same detection efficiency vs E curve. From former two advantages, we can know the most important advantages of the this system. Third advantage is economical system. The price of typical GM counter is about $1000. But the price of our system is below $100 because of plastic scintillator and simple electronics. The role of scintillation material is emitting scintillation which is the flash of light produced in certain materials when they absorb ionizing radiation. Plastic scintillator is organic scintillator which is kind of hydrocarbons. The special point are cheap price, large size production(∼ton), moderate light output, fast light emission(ns). And the role of Al filter is equalizing counting efficiency of air and scintillator for

  17. Radiation Damage in Scintillating Crystals

    CERN Document Server

    Zhu Ren Yuan

    1998-01-01

    Crystal Calorimetry in future high energy physics experiments faces a new challenge to maintain its precision in a hostile radiation environment. This paper discusses the effects of radiation damage in scintillating crystals, and concludes that the predominant radiation damage effect in crystal scintillators is the radiation induced absorption, or color center formation, not the loss of the scintillation light yield. The importance of maintaining crystal's light response uniformity and the feasibility to build a precision crystal calorimeter under radiation are elaborated. The mechanism of the radiation damage in scintillating crystals is also discussed. While the damage in alkali halides is found to be caused by the oxygen or hydroxyl contamination, it is the structure defects, such as oxygen vacancies, cause damage in oxides. Material analysis methods used to reach these conclusions are presented in details.

  18. CeF3(Ba) radiation hard scintillator for electromagnetic calorimeters

    International Nuclear Information System (INIS)

    Aseev, A.A.; Devitsin, E.G.; Kozlov, V.A.; Hovepyan, Yu.I.; Potashov, S.Yu.; Sokolovsky, K.A.; Uvarova, T.V.; Vasilchenko, V.G.

    1992-01-01

    The influence of divalent fluoride dopants BaF 2 , CaF 2 , SrF 2 on radiation and luminescent properties of CeF 3 crystal is studied. A high radiation hardness (>10 8 rad) has been obtained for CeF 3 crystals doped with BaF 2 . (orig.)

  19. Radiation hardness of LuAG:Ce and LuAG:Pr scintillator crystals

    CERN Document Server

    Derdzyan, M V; Belsky, A; Dujardin, C; Lecoq, P; Lucchini, M; Ovanesyan, K L; Pauwels, K; Pedrini, C; Petrosyan, A G

    2012-01-01

    Single crystals of LuAG:Ce, LuAG:Pr and un-doped LuAG were grown by the vertical Bridgman method and studied for radiation hardness under gamma-rays with doses in the range 10-10(5) Gy (Co-60). A wide absorption band peaking at around 600 nm springs up in all three types of crystals after the irradiations. The second band peaking at around 375 nm appears in both LuAG:Pr and un-doped LuAG. Compositional variations have been done to reveal the spectral behavior of induced color centers in more detail and to understand their origin. Similarities in behavior of Yb2+ centers in as-grown garnets are found, indicating that radiation induced color centers can be associated with residual trace amounts of Yb present in the raw materials. Un-doped LuAG and LuAG:Ce demonstrate moderate radiation hardness (the induced absorption coefficients being equal to 0.05-0.08 cm(-1) for accumulated doses of 10(3)-10(4) Gy), while LuAG:Pr is less radiation hard. The ways to improve the radiation hardness are discussed.

  20. Radiation damage studies on new liquid scintillators and liquid-core scintillating fibers

    International Nuclear Information System (INIS)

    Golovkin, S.V.

    1994-01-01

    The radiation resistant of some new liquid scintillation and capillaries filled with liquid scintillators has been presented. It was found that scintillation efficiency of the scintillator based on 1-methyl naphthalene with a new R39 only by 10% at the dose of 190 Mrad and the radiation resistance of thin liquid-core scintillating was decreased fibers exceeded 60 Mrad. 35 refs

  1. Improved Growth Methods for LaBr3 Scintillation Radiation Detectors

    International Nuclear Information System (INIS)

    McGregor, Douglas S.

    2011-01-01

    The objective is to develop advanced materials for deployment as high-resolution gamma ray detectors. Both LaBr3 and CeBr3 are advanced scintillation materials, and will be studied in this research. Prototype devices, in collaboration Sandia National Laboratories, will be demonstrated along with recommendations for mass production and deployment. It is anticipated that improved methods of crystal growth will yield larger single crystals of LaBr3 for deployable room-temperature operated gamma radiation spectrometers. The growth methods will be characterized. The LaBr3 and CeBr3 scintillation crystals will be characterized for light yield, spectral resolution, and for hardness.

  2. Polysiloxane scintillator composition

    Science.gov (United States)

    Walker, J.K.

    1992-05-05

    A plastic scintillator useful for detecting ionizing radiation comprising a matrix which comprises an optically transparent polysiloxane having incorporated therein at least one ionizing radiation-hard fluor capable of converting electromagnetic energy produced in the polysiloxane upon absorption of ionizing radiation to detectable light.

  3. A comparative study of the radiation hardness of plastic scintillators for the upgrade of the Tile Calorimeter of the ATLAS detector

    Science.gov (United States)

    Liao, S.; Erasmus, R.; Jivan, H.; Pelwan, C.; Peters, G.; Sideras-Haddad, E.

    2015-10-01

    The influence of radiation on the light transmittance of plastic scintillators was studied experimentally. The high optical transmittance property of plastic scintillators makes them essential in the effective functioning of the Tile calorimeter of the ATLAS detector at CERN. This significant role played by the scintillators makes this research imperative in the movement towards the upgrade of the tile calorimeter. The radiation damage of polyvinyl toluene (PVT) based plastic scintillators was studied, namely, EJ-200, EJ-208 and EJ-260, all manufactured and provided to us by ELJEN technology. In addition, in order to compare to scintillator brands actually in use at the ATLAS detector currently, two polystyrene (PS) based scintillators and an additional PVT based scintillator were also scrutinized in this study, namely, Dubna, Protvino and Bicron, respectively. All the samples were irradiated using a 6 MeV proton beam at different doses at iThemba LABS Gauteng. The radiation process was planned and mimicked by doing simulations using a SRIM program. In addition, transmission spectra for the irradiated and unirradiated samples of each grade were obtained, observed and analyzed.

  4. Evaluating scintillator performance in time-resolved hard X-ray studies at synchrotron light sources

    International Nuclear Information System (INIS)

    Rutherford, Michael E.; Chapman, David J.; White, Thomas G.; Drakopoulos, Michael; Rack, Alexander; Eakins, Daniel E.

    2016-01-01

    Scintillator performance in time-resolved, hard, indirect detection X-ray studies on the sub-microsecond timescale at synchrotron light sources is reviewed, modelled and examined experimentally. LYSO:Ce is found to be the only commercially available crystal suitable for these experiments. The short pulse duration, small effective source size and high flux of synchrotron radiation is ideally suited for probing a wide range of transient deformation processes in materials under extreme conditions. In this paper, the challenges of high-resolution time-resolved indirect X-ray detection are reviewed in the context of dynamic synchrotron experiments. In particular, the discussion is targeted at two-dimensional integrating detector methods, such as those focused on dynamic radiography and diffraction experiments. The response of a scintillator to periodic synchrotron X-ray excitation is modelled and validated against experimental data collected at the Diamond Light Source (DLS) and European Synchrotron Radiation Facility (ESRF). An upper bound on the dynamic range accessible in a time-resolved experiment for a given bunch separation is calculated for a range of scintillators. New bunch structures are suggested for DLS and ESRF using the highest-performing commercially available crystal LYSO:Ce, allowing time-resolved experiments with an interframe time of 189 ns and a maximum dynamic range of 98 (6.6 bits)

  5. Evaluating scintillator performance in time-resolved hard X-ray studies at synchrotron light sources

    Energy Technology Data Exchange (ETDEWEB)

    Rutherford, Michael E.; Chapman, David J.; White, Thomas G. [Imperial College London, London (United Kingdom); Drakopoulos, Michael [Diamond Light Source, I12 Joint Engineering, Environmental, Processing (JEEP) Beamline, Didcot, Oxfordshire (United Kingdom); Rack, Alexander [European Synchrotron Radiation Facility, Grenoble (France); Eakins, Daniel E., E-mail: d.eakins@imperial.ac.uk [Imperial College London, London (United Kingdom)

    2016-03-24

    Scintillator performance in time-resolved, hard, indirect detection X-ray studies on the sub-microsecond timescale at synchrotron light sources is reviewed, modelled and examined experimentally. LYSO:Ce is found to be the only commercially available crystal suitable for these experiments. The short pulse duration, small effective source size and high flux of synchrotron radiation is ideally suited for probing a wide range of transient deformation processes in materials under extreme conditions. In this paper, the challenges of high-resolution time-resolved indirect X-ray detection are reviewed in the context of dynamic synchrotron experiments. In particular, the discussion is targeted at two-dimensional integrating detector methods, such as those focused on dynamic radiography and diffraction experiments. The response of a scintillator to periodic synchrotron X-ray excitation is modelled and validated against experimental data collected at the Diamond Light Source (DLS) and European Synchrotron Radiation Facility (ESRF). An upper bound on the dynamic range accessible in a time-resolved experiment for a given bunch separation is calculated for a range of scintillators. New bunch structures are suggested for DLS and ESRF using the highest-performing commercially available crystal LYSO:Ce, allowing time-resolved experiments with an interframe time of 189 ns and a maximum dynamic range of 98 (6.6 bits)

  6. Radiation damage in plastic scintillators

    International Nuclear Information System (INIS)

    Majewski, S.

    1990-01-01

    Results of radiation damage studies in plastic scintillators are reviewed and critically analyzed from the point of view of applications of plastic scintillators in calorimetric detectors for the SSC. Damage to transmission and to fluorescent yield in different conditions is discussed. New directions in R ampersand D are outlined. Several examples are given of the most recent data on the new scintillating materials made with old and new plastics and fluors, which are exhibiting significantly improved radiation resistance. With a present rate of a vigorous R D programme, the survival limits in the vicinity of 100 MRad seem to be feasible within a couple of years

  7. Recovery and permanent radiation damage of plastic scintillators at different dose rates

    International Nuclear Information System (INIS)

    Bicken, B.; Holm, U.; Marckmann, T.; Wick, K.; Rhode, M.

    1990-01-01

    This paper reports on the radiation stability of plastic scintillators and wavelength shifters for the calorimeter of the ZEUS detector by irradiating them with protons, a 60 Co-source, and depleted uranium. Changes in light yield, absorption length and absorption coefficient have been measured for storage in inert and oxygen atmospheres during and after irradiation. Radiation doses up to 40 kGy with dose rates of 30 up to 2000 Gy/h have been applied. The polystyrene based scintillator SCSN-38 and the wavelength shifters Y-7 and K-27 in PMMA show an additional absorption but a recovery in air to a low permanent damage (at 10 kGy) which is proportional to the applied dose. Series investigations on samples of all production cycles of the ZEUS scintillators with high dose rates show only minor differences in radiation hardness. The recovery is described by a simple oxygen diffusion model for high and medium dose rates down to 30 Gy/h. During long term irradiations at low dose rates (<100 Gy/h) of 3 mm thick SCSN-38 in air the radiation damage recovers to a permanent damage which does not depend on the dose rate. On the other hand the radiation damage at very low dose rates (17 Gy/a) seems to be higher than expected for the accumulated dose

  8. Radiation effects in polymers for plastic scintillation detectors

    International Nuclear Information System (INIS)

    Pla-Dalmau, A.; Bross, A.D.; Hurlbut, C.R.; Moser, S.W.

    1994-01-01

    Recent developments in both scintillating plastic optical fibers and photon detection devices have spawned new applications for plastic scintillator detectors. This renewed attention has encouraged research that addresses the radiation stability of plastic scintillators. The optical quality of the polymer degrades with exposure to ionizing radiation and thus the light yield of the detector decreases. A complete understanding of all the mechanisms contributing to this radiation-induced degradation of the polymer can lead to techniques that will extend the radiation stability of these materials. Various radiation damage studies have been performed under different atmospheres and dose rates. Currently, the use of additives to preserve the optical properties of the polymer matrix under radiation is being investigated. The authors discuss the effect of certain antioxidants, plasticizers, and cross-linking agents on the radiation resilience of plastic scintillators

  9. Waveshifters and Scintillators for Ionizing Radiation Detection

    International Nuclear Information System (INIS)

    Baumgaugh, B.; Bishop, J.; Karmgard, D.; Marchant, J.; McKenna, M.; Ruchti, R.; Vigneault, M.; Hernandez, L.; Hurlbut, C.

    2007-01-01

    Scintillation and waveshifter materials have been developed for the detection of ionizing radiation in an STTR program between Ludlum Measurements, Inc. and the University of Notre Dame. Several new waveshifter materials have been developed which are comparable in efficiency and faster in fluorescence decay than the standard material Y11 (K27) used in particle physics for several decades. Additionally, new scintillation materials useful for fiber tracking have been developed which have been compared to 3HF. Lastly, work was done on developing liquid scintillators and paint-on scintillators and waveshifters for high radiation environments

  10. Radiation hardness qualification of PbWO4 scintillation crystals for the CMS Electromagnetic Calorimeter

    CERN Document Server

    Adzic, P.; Andelin, D.; Anicin, I.; Antunovic, Z.; Arcidiacono, R.; Arenton, M.W.; Auffray, E.; Argiro, S.; Askew, A.; Baccaro, S.; Baffioni, S.; Balazs, M.; Bandurin, D.; Barney, D.; Barone, L.M.; Bartoloni, A.; Baty, C.; Beauceron, S.; Bell, K.W.; Bernet, C.; Besancon, M.; Betev, B.; Beuselinck, R.; Biino, C.; Blaha, J.; Bloch, P.; Borisevitch, A.; Bornheim, A.; Bourotte, J.; Brown, R.M.; Buehler, M.; Busson, P.; Camanzi, B.; Camporesi, T.; Cartiglia, N.; Cavallari, F.; Cecilia, A.; Chang, P.; Chang, Y.H.; Charlot, C.; Chen, E.A.; Chen, W.T.; Chen, Z.; Chipaux, R.; Choudhary, B.C.; Choudhury, R.K.; Cockerill, D.J.A.; Conetti, S.; Cooper, S.I.; Cossutti, F.; Cox, B.; Cussans, D.G.; Dafinei, I.; Da Silva Di Calafiori, D.R.; Daskalakis, G.; David, A.; Deiters, K.; Dejardin, M.; De Benedetti, A.; Della Ricca, G.; Del Re, D.; Denegri, D.; Depasse, P.; Descamps, J.; Diemoz, M.; Di Marco, E.; Dissertori, G.; Dittmar, M.; Djambazov, L.; Djordjevic, M.; Dobrzynski, L.; Dolgopolov, A.; Drndarevic, S.; Drobychev, G.; Dutta, D.; Dzelalija, M.; Elliott-Peisert, A.; El Mamouni, H.; Evangelou, I.; Fabbro, B.; Faure, J.L.; Fay, J.; Fedorov, A.; Ferri, F.; Franci, D.; Franzoni, G.; Freudenreich, K.; Funk, W.; Ganjour, S.; Gascon, S.; Gataullin, M.; Gentit, F.X.; Ghezzi, A.; Givernaud, A.; Gninenko, S.; Go, A.; Gobbo, B.; Godinovic, N.; Golubev, N.; Govoni, P.; Grant, N.; Gras, P.; Haguenauer, M.; Hamel de Monchenault, G.; Hansen, M.; Haupt, J.; Heath, H.F.; Heltsley, B.; Cornell U., LNS.; Hintz, W.; Hirosky, R.; Hobson, P.R.; Honma, A.; Hou, G.W.S.; Hsiung, Y.; Huhtinen, M.; Ille, B.; Ingram, Q.; Inyakin, A.; Jarry, P.; Jessop, C.; Jovanovic, D.; Kaadze, K.; Kachanov, V.; Kailas, S.; Kataria, S.K.; Kennedy, B.W.; Kokkas, P.; Kolberg, T.; Korjik, M.; Krasnikov, N.; Krpic, D.; Kubota, Y.; Kuo, C.M.; Kyberd, P.; Kyriakis, A.; Lebeau, M.; Lecomte, P.; Lecoq, P.; Ledovskoy, A.; Lethuillier, M.; Lin, S.W.; Lin, W.; Litvine, V.; Locci, E.; Longo, E.; Loukas, D.; Luckey, P.D.; Lustermann, W.; Ma, Y.; Malberti, M.; Malcles, J.; Maletic, D.; Manthos, N.; Maravin, Y.; Marchica, C.; Marinelli, N.; Markou, A.; Markou, C.; Marone, M.; Matveev, V.; Mavrommatis, C.; Meridiani, P.; Milenovic, P.; Mine, P.; Missevitch, O.; Mohanty, A.K.; Moortgat, F.; Musella, P.; Musienko, Y.; Nardulli, A.; Nash, J.; Nedelec, P.; Negri, P.; Newman, H.B.; Nikitenko, A.; Nessi-Tedaldi, F.; Obertino, M.M.; Organtini, G.; Orimoto, T.; Paganoni, M.; Paganini, P.; Palma, A.; Pant, L.; Papadakis, A.; Papadakis, I.; Papadopoulos, I.; Paramatti, R.; Parracho, P.; Pastrone, N.; Patterson, J.R.; Pauss, F.; Peigneux, J.P.; Petrakou, E.; Phillips, D.G.; Piroue, P.; Ptochos, F.; Puljak, I.; Pullia, A.; Punz, T.; Puzovic, J.; Ragazzi, S.; Rahatlou, S.; Rander, J.; Razis, P.A.; Redaelli, N.; Renker, D.; Reucroft, S.; Ribeiro, P.; Rogan, C.; Ronquest, M.; Rosowsky, A.; Rovelli, C.; Rumerio, P.; Rusack, R.; Rusakov, S.V.; Ryan, M.J.; Sala, L.; Salerno, R.; Schneegans, M.; Seez, C.; Sharp, P.; Shepherd-Themistocleous, C.H.; Shiu, J.G.; Shivpuri, R.K.; Shukla, P.; Siamitros, C.; Sillou, D.; Silva, J.; Silva, P.; Singovsky, A.; Sirois, Y.; Sirunyan, A.; Smith, V.J.; Stockli, F.; Swain, J.; Tabarelli de Fatis, T.; Takahashi, M.; Tancini, V.; Teller, O.; Theofilatos, K.; Thiebaux, C.; Timciuc, V.; Timlin, C.; Titov, Maxim P.; Topkar, A.; Triantis, F.A.; Troshin, S.; Tyurin, N.; Ueno, K.; Uzunian, A.; Varela, J.; Verrecchia, P.; Veverka, J.; Virdee, T.; Wang, M.; Wardrope, D.; Weber, M.; Weng, J.; Williams, J.H.; Yang, Y.; Yaselli, I.; Yohay, R.; Zabi, A.; Zelepoukine, S.; Zhang, J.; Zhang, L.Y.; Zhu, K.; Zhu, R.Y.

    2010-01-01

    Ensuring the radiation hardness of PbWO4 crystals was one of the main priorities during the construction of the electromagnetic calorimeter of the CMS experiment at CERN. The production on an industrial scale of radiation hard crystals and their certification over a period of several years represented a difficult challenge both for CMS and for the crystal suppliers. The present article reviews the related scientific and technological problems encountered.

  11. Report on radiation exposure of lead-scintillator stack

    International Nuclear Information System (INIS)

    Underwood, D.G.

    1990-01-01

    A stack of lead and scintillator was placed in a neutral beam obtained from targeting 800 GeV protons. Small pieces of film containing radiochromic dye were placed adjacent to the layers of scintillator for the purpose of measuring the radiation dose to the scintillator. Our motivation was to calibrate the radiation dose obtainable in this manner for future tests of scintillator for SSC experiments and to relate dose to flux to check absolute normalization for calculations. We also observed several other radiation effects which should be considered for both damage and compensation in a calorimeter

  12. Radiation Hardness Study of CsI(Tl) Crystals for Belle II Calorimeter

    CERN Document Server

    Matvienko, D V; Sedov, E V; Shwartz, B A

    2017-01-01

    The Belle II calorimeter (at least, its barrel part) consists of CsI(Tl) scintillation crystals which have been used at the Belle experiment. We perform the radiation hardness study of some typical Belle crystals and conclude their light output reductions are acceptable for Belle II experiment where the absorption dose can reach 10 krad during the detector operation. CsI(Tl) crystals have high stablity and low maintenance cost and are considered as possible option for the calorimeter of the future Super-Charm-Tau factory (SCT) in Novosibirsk. Our study demonstrates sufficiently high radiation hardness of CsI(Tl) crystals for SCT conditions.

  13. Semiconductor high-energy radiation scintillation detector

    International Nuclear Information System (INIS)

    Kastalsky, A.; Luryi, S.; Spivak, B.

    2006-01-01

    We propose a new scintillation-type detector in which high-energy radiation generates electron-hole pairs in a direct-gap semiconductor material that subsequently recombine producing infrared light to be registered by a photo-detector. The key issue is how to make the semiconductor essentially transparent to its own infrared light, so that photons generated deep inside the semiconductor could reach its surface without tangible attenuation. We discuss two ways to accomplish this, one based on doping the semiconductor with shallow impurities of one polarity type, preferably donors, the other by heterostructure bandgap engineering. The proposed semiconductor scintillator combines the best properties of currently existing radiation detectors and can be used for both simple radiation monitoring, like a Geiger counter, and for high-resolution spectrography of the high-energy radiation. An important advantage of the proposed detector is its fast response time, about 1 ns, essentially limited only by the recombination time of minority carriers. Notably, the fast response comes without any degradation in brightness. When the scintillator is implemented in a qualified semiconductor material (such as InP or GaAs), the photo-detector and associated circuits can be epitaxially integrated on the scintillator slab and the structure can be stacked-up to achieve virtually any desired absorption capability

  14. Methacrylate based cross-linkers for improved thermomechanical properties and retention of radiation detection response in plastic scintillators

    Science.gov (United States)

    Mahl, Adam; Lim, Allison; Latta, Joseph; Yemam, Henok A.; Greife, Uwe; Sellinger, Alan

    2018-03-01

    Pulse shape discrimination (PSD) is an important method that can efficiently sort and separate neutron and gamma radiation signals. PSD is currently achieved in plastic scintillators by over-doping poly(vinyl toluene) (PVT) matrices with fluorescent molecules. Meaningful separation of the signals requires addition of >20 wt% 2,5-diphenyloxazole (PPO) fluor in PVT. At these concentrations PPO acts as a plasticizer, negatively affecting the physical properties of the final plastic such as hardness, machinability, and thermomechanical stability. This work addresses these issues by implementing a cost-effective solution using cross-linking chemistry via commercially available bisphenol A dimethacrylate (BPA-DM), and a synthesized fluorinated analogue. Both improve the physical properties of over-doped PPO based plastic scintillators without degrading the measured light yield or PSD and Figure of Merit (FoM). In addition, the fluorinated analogue appears to enhance the hydrophobicity of the surface of the plastic scintillators, which may improve the scintillators' resistance to water diffusion and subsequent radiation response degradation. The new formulations improve the feasibility of widely deploying long lifetime PSD capable plastic scintillators in large area coverage assemblies.

  15. Long-distance transmission of light in a scintillator-based radiation detector

    Science.gov (United States)

    Dowell, Jonathan L.; Talbott, Dale V.; Hehlen, Markus P.

    2017-07-11

    Scintillator-based radiation detectors capable of transmitting light indicating the presence of radiation for long distances are disclosed herein. A radiation detector can include a scintillator layer and a light-guide layer. The scintillator layer is configured to produce light upon receiving incident radiation. The light-guide layer is configured to receive light produced by the scintillator layer and either propagate the received light through the radiation detector or absorb the received light and emit light, through fluorescence, that is propagated through the radiation detector. A radiation detector can also include an outer layer partially surrounding the scintillator layer and light-guide layer. The index of refraction of the light-guide layer can be greater than the index of refraction of adjacent layers.

  16. Fiber-optic multipoint radiation sensing system using waveguide scintillators

    International Nuclear Information System (INIS)

    Maekawa, Tatsuyuki; Yoda, Masaki; Tanaka, Koutarou; Masumaru, Tarou; Morimoto, Souichirou.

    1996-01-01

    Novel fiber-optic radiation sensors and a multipoint measurement method that takes advantage of them have been developed. The new sensor design, which we call a 'waveguide scintillator', consists of a scintillating material and a wavelength-shifting fiber (WLSF). The WLSF is embedded in the scintillating material, and each end is connected to a transparent optical fiber. These waveguide scintillators can be connected in series along an optical fiber loop to form a radiation monitoring system, and each end of the fiber loop is terminated with a photodetector. This new radiation monitoring arrangement dispenses with the need for electronic apparatus at each measuring point and consequently improves resistance to noise. Furthermore, it offers the advantages of multipoint monitoring - meaning that radiation intensity can be measured at multiple sensors - using only two photodetectors. We have examined the light output characteristics and time resolution of a prototype arrangement of these new waveguide scintillators, thus confirming the feasibility of multipoint measurements using a system of multiple waveguide scintillators connected in series in an optical fiber loop. (author)

  17. Composite scintillators for detection of ionizing radiation

    Science.gov (United States)

    Dai, Sheng [Knoxville, TN; Stephan, Andrew Curtis [Knoxville, TN; Brown, Suree S [Knoxville, TN; Wallace, Steven A [Knoxville, TN; Rondinone, Adam J [Knoxville, TN

    2010-12-28

    Applicant's present invention is a composite scintillator having enhanced transparency for detecting ionizing radiation comprising a material having optical transparency wherein said material comprises nano-sized objects having a size in at least one dimension that is less than the wavelength of light emitted by the composite scintillator wherein the composite scintillator is designed to have selected properties suitable for a particular application.

  18. Effects of radiation on scintillating fiber performance

    International Nuclear Information System (INIS)

    Bauer, M.L.; Cohn, H.; Efremenko, Yu.; Gordeev, A.; Kamyshkov, Yu.; Onopienko, D.; Savin, S.; Shmakov, K.; Tarkovsky, E.; Young, K.G.; Carey, R.; Rothman, M.; Sulak, L.; Worstell, W.; Parr, H.

    1992-01-01

    Continued rapid improvements in formulations for scintillating fibers require the ability to parameterize and predict effects of radiation on detector performance. Experimental techniques necessary to obtain needed information and calculational procedures used in performing predications for hadron scintillating fiber calorimetry in the Superconducting Supercollider environment are described. The experimental techniques involve control of the testing environment, consideration of dose rate effects, and other factors. These calculations involve the behavior of particle showers in the detector, expected levels of radiation, and parameterization of the radiation effects. A summary of significant work is also presented

  19. Effects of radiation on scintillating fiber performance

    International Nuclear Information System (INIS)

    Young, K.G.; Bauer, M.L.; Cohn, H.; Efremenko, Yu.; Gordeev, A.; Kamyshkov, Yu.; Onopienko, D.; Savin, S.; Shmakov, K.; Tarkovsky, E.; Carey, R.; Rothman, M.; Sulak, L.; Worstell, W.; Paar, H.

    1993-01-01

    Continued rapid improvements in formulations for scintillating fibers require the ability to parameterize and predict effects of radiation on detector performance. Experimental techniques necessary to obtain desired information and calculational procedures used in performing predictions for hadron scintillating fiber calorimetry in the Superconducting Supercollider environment are described. The experimental techniques involve control of the testing environment, consideration of dose rate effects, and other factors. The calculations involve the behavior of particle showers in the detector, expected levels of radiation, and parameterization of the radiation effects. A summary of significant work is also presented

  20. Radiation Damage Mechanism in PbWO4 Crystal and Radiation Hardness Quality Control of PWO Scintillators for CMS

    CERN Document Server

    Baccaro, Stefania; Borgia, Bruno; Cavallari, Francesca; Cecilia, Angelica; Dafinei, Ioan; Diemoz, Marcella; Lecoq, Paul; Longo, Egidio; Montecchi, Marco; Organtini, Giovanni; Salvatori, S

    1997-01-01

    The optical damage induced by UV light in PbWO4 crystals is found to be similar to that induced by g radiation. Due to the peculiarities of optical absorption in PbWO4, the damage induced by UV light is a bulk process. This fact has important consequences for the approach to be adopted both for the use of the crystal as scintillator and for the qualification methods foreseen in the Regional Centres of the ECAL CMS Collaboration.

  1. Silica scintillating materials prepared by sol-gel methods

    International Nuclear Information System (INIS)

    Werst, D.W.; Sauer, M.C. Jr.; Cromack, K.R.; Lin, Y.; Tartakovsky, E.A.; Trifunac, A.D.

    1993-01-01

    Silica was investigated as a rad-hard alternative to organic polymer hosts for organic scintillators. Silica sol-gels were prepared by hydrolysis of tetramethoxysilane in alcohol solutions. organic dyes were incorporated into the gels by dissolving in methanol at the sol stage of gel formation. The silica sol-gel matrix is very rad-hard. The radiation stability of silica scintillators prepared by this method is dye-limited. Transient radioluminescence was measured following excitation with 30 ps pulses of 20 MeV electrons

  2. Preparation and characterisation of radiation hard PbWO4 crystal scintillator

    International Nuclear Information System (INIS)

    Sabharwal, S.C.; Desai, D.G.; Sangeeta; Karandikar, S.C.; Chauhan, A.K.; Sangiri, A.K.; Keshwani, K.S.; Ahuja, M.N.

    1996-01-01

    The selective loss of one of the crystal constituents is found to be responsible for the yellowish coloration of PbWO 4 crystals. However, using the already pulled crystals as the starting charge for the subsequent growth, colorless crystals can be grown. The crystals exhibiting excellent transmission characteristics have been grown employing a low temperature gradient, a moderate rotation rate of 15 rpm and a pull speed of 1 mm/h. The colored crystals show some radiation damage on gamma irradiation, while the colorless ones remain unaffected even for irradiation doses as high as 10 Mrad. Both the types of crystals show the presence of weak thermoluminescence (TL) emission when high irradiation doses (similar 10 Mrad) are given. Only one TL glow peak is obtained in both the cases but the peak temperatures are different. The emission centers responsible for the TL emission are found to be the ones which give rise to the scintillation emission in the crystal. (orig.)

  3. Radiation-induced chemical processes in polystyrene scintillators

    International Nuclear Information System (INIS)

    Milinchuk, V.K.; Bolbit, N.M.; Klinshpont, E.R.; Tupikov, V.I.; Zhdanov, G.S.; Taraban, S.B.; Shelukhov, I.P.; Smoljanskii, A.S.

    1999-01-01

    The regularities established for macroradical accumulation and intensity of radioluminescence under γ-irradiation of a polystyrene scintillator prove benzyl macroradicals to be efficient quenchers of the excited scintillator molecules. Dissolved oxygen was determined to have a constant of the quenching rate 100 times lower than that of macroradicals. Oxygen is an efficient antirad because of participating in oxidation reactions and subsequent recombination of macroradicals. The method was developed to obtain a polymeric scintillator with a polystyrene matrix containing a dispersed system of pores and channels. Radiation resistance of such a scintillator is 5-10 times higher than that of standard types

  4. Detection of gamma-neutron radiation by solid-state scintillation detectors. Detection of gamma-neutron radiation by novel solid-state scintillation detectors

    Energy Technology Data Exchange (ETDEWEB)

    Ryzhikov, V.; Grinyov, B.; Piven, L.; Onyshchenko, G.; Sidletskiy, O. [Institute for Scintillation Materials of the NAS of Ukraine, Kharkov, (Ukraine); Naydenov, S. [Institute for Single Crystals of the National Academy of Sciences of Ukraine, Kharkov, (Ukraine); Pochet, T. [DETEC-Europe, Vannes (France); Smith, C. [Naval Postgraduate School, Monterey, CA (United States)

    2015-07-01

    It is known that solid-state scintillators can be used for detection of both gamma radiation and neutron flux. In the past, neutron detection efficiencies of such solid-state scintillators did not exceed 5-7%. At the same time it is known that the detection efficiency of the gamma-neutron radiation characteristic of nuclear fissionable materials is by an order of magnitude higher than the efficiency of detection of neutron fluxes alone. Thus, an important objective is the creation of detection systems that are both highly efficient in gamma-neutron detection and also capable of exhibiting high gamma suppression for use in the role of detection of neutron radiation. In this work, we present the results of our experimental and theoretical studies on the detection efficiency of fast neutrons from a {sup 239}Pu-Be source by the heavy oxide scintillators BGO, GSO, CWO and ZWO, as well as ZnSe(Te, O). The most probable mechanism of fast neutron interaction with nuclei of heavy oxide scintillators is the inelastic scattering (n, n'γ) reaction. In our work, fast neutron detection efficiencies were determined by the method of internal counting of gamma-quanta that emerge in the scintillator from (n, n''γ) reactions on scintillator nuclei with the resulting gamma energies of ∼20-300 keV. The measured efficiency of neutron detection for the scintillation crystals we considered was ∼40-50 %. The present work included a detailed analysis of detection efficiency as a function of detector and area of the working surface, as well as a search for new ways to create larger-sized detectors of lower cost. As a result of our studies, we have found an unusual dependence of fast neutron detection efficiency upon thickness of the oxide scintillators. An explanation for this anomaly may involve the competition of two factors that accompany inelastic scattering on the heavy atomic nuclei. The transformation of the energy spectrum of neutrons involved in the (n, n

  5. Plastic scintillation dosimetry for radiation therapy: minimizing capture of Cerenkov radiation noise

    International Nuclear Information System (INIS)

    Beddar, A Sam; Suchowerska, Natalka; Law, Susan H

    2004-01-01

    Over the last decade, there has been an increased interest in scintillation dosimetry using small water-equivalent plastic scintillators, because of their favourable characteristics when compared with other more commonly used detector systems. Although plastic scintillators have been shown to have many desirable dosimetric properties, as yet there is no successful commercial detector system of this type available for routine clinical use in radiation oncology. The main factor preventing this new technology from realizing its full potential in commercial applications is the maximization of signal coupling efficiency and the minimization of noise capture. A principal constituent of noise is Cerenkov radiation. This study reports the calculated capture of Cerenkov radiation by an optical fibre in the special case where the radiation is generated by a relativistic particle on the fibre axis and the fibre axis is parallel to the Cerenkov cone. The fraction of radiation captured is calculated as a function of the fibre core refractive index and the refractive index difference between the core and the cladding of the fibre for relativistic particles. This is then used to deduce the relative intensity captured for a range of fibre core refractive indices and fibre core-cladding refractive index differences. It is shown that the core refractive index has little effect on the amount of radiation captured compared to the refractive index difference. The implications of this result for the design of radiation therapy plastic scintillation dosimeters are considered

  6. A novel epitaxially grown LSO-based thin-film scintillator for micro-imaging using hard synchrotron radiation

    Energy Technology Data Exchange (ETDEWEB)

    Douissard, P.A.; Martin, T.; Chevalier, V.; Rack, A. [European Synchrotron Radiat Facil, F-38043 Grenoble, (France); Cecilia, A.; Baumbach, T.; Rack, A. [Karlsruhe Inst Technol ANKA, D-76021 Karlsruhe, (Germany); Couchaud, M. [CEA LETI, F-38054 Grenoble, (France); Dupre, K. [FEE GmbH, D-55743 Idar Oberstein, (Germany); Kuhbacher, M. [Helmholtz Zentrum Berlin Mat and Energie, D-14109 Berlin, (Germany)

    2010-07-01

    The efficiency of high-resolution pixel detectors for hard X-rays is nowadays one of the major criteria which drives the feasibility of imaging experiments and in general the performance of an experimental station for synchrotron-based microtomography and radiography. Here the luminescent screen used for the indirect detection is focused on in order to increase the detective quantum efficiency a novel scintillator based on doped Lu{sub 2}SiO{sub 5} (LSO), epitaxially grown as thin film via the liquid phase epitaxy technique. It is shown that, by using adapted growth and doping parameters as well as a dedicated substrate, the scintillation behaviour of a LSO-based thin crystal together with the high stopping power of the material allows for high-performance indirect X-ray detection. In detail, the conversion efficiency, the radioluminescence spectra, the optical absorption spectra under UV/visible-light and the afterglow are investigated. A set-up to study the effect of the thin-film scintillator's temperature on its conversion efficiency is described as well it delivers knowledge which is important when working with higher photon flux densities and the corresponding high heat load on the material. Additionally, X-ray imaging systems based on different diffraction-limited visible-light optics and CCD cameras using among others LSO-based thin film are compared. Finally, the performance of the LSO thin film is illustrated by imaging a honey bee leg, demonstrating the value of efficient high-resolution computed tomography for life sciences. (authors)

  7. Development of radiation detection and measurement system - Development of scintillation radiation sensor

    Energy Technology Data Exchange (ETDEWEB)

    Kang, Hee Dong; Kim, Wan [Kyungpook National University, Taegu (Korea); Kim, Do Sung [Taegu University, Taegu (Korea)

    2000-03-01

    We have been fabricated CsI(Tl) scintillation crystals and plastic scintillators for radiation-based measuring equipment. CsI (Tl) single crystals doped with thallium as an activator were grown using the Czochralski method. The crystal structure of grown CsI(Tl) was bcc, and it was confirmed that its lattice constant was 4,568 A. The spectral range of luminescence of CsI(Tl) was 350 {approx} 700 nm independent of thallium concentration, and the fast component of the luminescence was decreased with increasing thallium concentration. The energy resolution of CsI(Tl) scintillator doped with 0.1 mole% thallium was about 9% for 137 Cs {gamma}-rays. The relation formula of {gamma}-ray energy versus energy resolution was ln(FWHM%)=-0.705ln({epsilon})+6.75. The radiation damage of CsI(Tl) increased in proportion to thallium concentration and radiation damage of CsI(Tl) increased in proportion to thallium concentration and radiation dosage, and the irradiated crystals were colored reddish. The radiation induced absorption bands appeared around 355, 425, 520 and 555 nm, and their energy level were about 3.50, 2.88, 2.39 and 2.21 eV. Plastic scintillators were fabricated thermal polymerization method. Those were polymerizing at 120 deg. C, during 72 hours, and annealing at 75 deg. C, during 24 hours. When the concentration of 1st solute was 1.5 wt% and concentration of 2nd solute was 0.01 wt%, the characteristics of scintillation were very excellent. Also 3.0 wt% tetraphenyl lead were loaded to improve the detection efficiency of {gamma}-ray. The range of emission spectrum was 400 {approx} 450nm, and the central peak was 415 nm. The radiation damage was not appear under 1*10{sup 3}Gy, but the color of plastic scintillator was changed to brown, over 1*10{sup 4}Gy exposured. 84 refs., 39 figs. (Author)

  8. Factors determining radiation stability of plastic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Barashkov, N.N. [Texas Univ., Richardson, TX (United States). Dept. of Chemistry; Gunder, O.A.; Voronkina, N.I. [National Ukrainian Academy of Science, Kharkov (Ukraine). Inst. for Single Crystals; Milinchuk, V.K. [Karpov Inst. of Physical Chemistry, Moscow (Russian Federation)

    1996-11-01

    Polystyrene (PS) and polyvinylxylene (PVX) are the base materials for plastic scintillators u.v.-Vis spectrophotometry, luminescence and ESR spectroscopy were performed on irradiated samples of PS and PVX with the p-terphenyl (TP) as the primary luminophore and 1,4-di-2(5-phenyloxazolyl-1,3) benzene (POPOP) as the secondary one. Dependence of the radioluminescence intensity of the scintillators on the concentration of the macroradicals formed by irradiation was investigated. Dose dependence of the radiation stability of scintillators is discussed. (author).

  9. Factors determining radiation stability of plastic scintillators

    International Nuclear Information System (INIS)

    Barashkov, N.N.

    1996-01-01

    Polystyrene (PS) and polyvinylxylene (PVX) are the base materials for plastic scintillators u.v.-Vis spectrophotometry, luminescence and ESR spectroscopy were performed on irradiated samples of PS and PVX with the p-terphenyl (TP) as the primary luminophore and 1,4-di-2(5-phenyloxazolyl-1,3) benzene (POPOP) as the secondary one. Dependence of the radioluminescence intensity of the scintillators on the concentration of the macroradicals formed by irradiation was investigated. Dose dependence of the radiation stability of scintillators is discussed. (author)

  10. Radiation converter scintillator screen and its manufacturing process

    International Nuclear Information System (INIS)

    Delattre, D.; Rougeot, H.; Tassin, C.

    1984-01-01

    The present invention concerns scintillating screens receiving X or gamma radiation and converting it in luminous photons. The screen comprises a needle structure scintillating material. Its concave surface is quite smooth. The screen is obtained by evaporation on a frame having a perfectly smooth convex surface; the constituting material has a thermal dilatation coefficient different from the scintillating material one. After evaporation, the scintillating screen is set apart from the frame by simple heating. It is used for radiological image intensifier tubes and scintigraphy tubes [fr

  11. Factors determining radiation stability of plastic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Gunder, O.A.; Voronkina, N.I. [National Ukranian Academy of Science, Kharkov (Ukraine). Inst. for Single Crystals; Barashkov, N.N.; Milinchuk, V.K.; Jdanov, G.S. [Karpov Inst. of Physical Chemistry, Moscow (Russian Federation)

    1995-07-01

    Polystyrene (PS) and polyvinylxylene (PVX) are the base materials for plastic scintillators. UV-Vis spectrophotometry, luminescence and EPR spectroscopy were performed on irradiated samples of PS and PVX with the p-terphenyl (TP) as primary luminophore and 1,4-di-2(5-phenyloxazolyl-1,3)benzene (POPOP) as secondary one. Dependence of the radioluminescence intensity of the scintillators on the concentration of the macroradicals formed in them in the process of irradiation was investigated. Dose dependence of the radiation stability of scintillators is discussed. (Author).

  12. Factors determining radiation stability of plastic scintillators

    Science.gov (United States)

    Gunder, O. A.; Voronkina, N. I.; Barashkov, N. N.; Milinchuk, V. K.; Jdanov, G. S.

    1995-07-01

    Polystyrene (PS) and polyvinylxylene (PVX) are the base materials for plastic scintillators. UV-Vis spectrophotometry, luminescence and EPR spectroscopy were performed on irradiated samples of PS and PVX with the p-terphenyl (TP) as primary luminophore and 1,4-di-2(5-phenyloxazolyl-1,3)benzene (POPOP) as secondary one. Dependence of the radioluminescence intensity of the scintillators on the concentration of the macroradicals formed in them in the process of irradiation was investigated. Dose dependence of the radiation stability of scintillators is discussed.

  13. Factors determining radiation stability of plastic scintillators

    International Nuclear Information System (INIS)

    Gunder, O.A.; Voronkina, N.I.

    1995-01-01

    Polystyrene (PS) and polyvinylxylene (PVX) are the base materials for plastic scintillators. UV-Vis spectrophotometry, luminescence and EPR spectroscopy were performed on irradiated samples of PS and PVX with the p-terphenyl (TP) as primary luminophore and 1,4-di-2(5-phenyloxazolyl-1,3)benzene (POPOP) as secondary one. Dependence of the radioluminescence intensity of the scintillators on the concentration of the macroradicals formed in them in the process of irradiation was investigated. Dose dependence of the radiation stability of scintillators is discussed. (Author)

  14. Quenching the scintillation in CF4 Cherenkov gas radiator

    International Nuclear Information System (INIS)

    Blake, T.; D'Ambrosio, C.; Easo, S.; Eisenhardt, S.; Fitzpatrick, C.; Forty, R.; Frei, C.; Gibson, V.; Gys, T.; Harnew, N.; Hunt, P.; Jones, C.R.; Lambert, R.W.; Matteuzzi, C.; Muheim, F.; Papanestis, A.; Perego, D.L.; Piedigrossi, D.; Plackett, R.; Powell, A.

    2015-01-01

    CF 4 is used as a Cherenkov gas radiator in one of the Ring Imaging Cherenkov detectors at the LHCb experiment at the CERN Large Hadron Collider. CF 4 is well known to have a high scintillation photon yield in the near and far VUV, UV and in the visible wavelength range. A large flux of scintillation photons in our photon detection acceptance between 200 and 800 nm could compromise the particle identification efficiency. We will show that this scintillation photon emission system can be effectively quenched, consistent with radiationless transitions, with no significant impact on the photons resulting from Cherenkov radiation

  15. Radiation-damage measurements on PVT-based plastic scintillators

    International Nuclear Information System (INIS)

    Ilie, S.; Schoenbacher, H.; Tavlet, M.

    1993-01-01

    Samples of PVT-based plastic scintillators produced by Nuclear Enterprise Technology Ltd. (NET) were irradiated up to 9 kGy, both with a gamma source and within a typical accelerator radiation field (CERN PS ACOL Irradiation Facility). The consequent reduction of scintillating efficiency and light transmission were measured, as well as subsequent recovery, over a period of several months. The main results show that irradiation affects more the light transmission than the light emission. The radiation type does not affect either the amount of transmission reduction or the recovery. Observations were also made by means of polarized light. Non-uniformities and internal stresses were observed in scintillator bulks which were polymerized too quickly. These defects influence the light transmission. (orig.)

  16. Conference on Engineering of Scintillation Materials and Radiation Technologies

    CERN Document Server

    Gektin, Alexander

    2017-01-01

    This volume provides a broad overview of the latest achievements in scintillator development, from theory to applications, and aiming for a deeper understanding of fundamental processes, as well as the discovery and availability of components for the production of new generations of scintillation materials. It includes papers on the microtheory of scintillation and the initial phase of luminescence development, applications of the various materials, and development and characterization of ionizing radiation detection equipment. The book also touches upon the increased demand for cryogenic scintillators, the renaissance of  garnet materials for scintillator applications, nano-structuring in scintillator development, development and applications for security, and exploration of hydrocarbons and ecological monitoring.

  17. Seeing the invisible: direct visualization of therapeutic radiation beams using air scintillation.

    Science.gov (United States)

    Fahimian, Benjamin; Ceballos, Andrew; Türkcan, Silvan; Kapp, Daniel S; Pratx, Guillem

    2014-01-01

    To assess whether air scintillation produced during standard radiation treatments can be visualized and used to monitor a beam in a nonperturbing manner. Air scintillation is caused by the excitation of nitrogen gas by ionizing radiation. This weak emission occurs predominantly in the 300-430 nm range. An electron-multiplication charge-coupled device camera, outfitted with an f/0.95 lens, was used to capture air scintillation produced by kilovoltage photon beams and megavoltage electron beams used in radiation therapy. The treatment rooms were prepared to block background light and a short-pass filter was utilized to block light above 440 nm. Air scintillation from an orthovoltage unit (50 kVp, 30 mA) was visualized with a relatively short exposure time (10 s) and showed an inverse falloff (r(2) = 0.89). Electron beams were also imaged. For a fixed exposure time (100 s), air scintillation was proportional to dose rate (r(2) = 0.9998). As energy increased, the divergence of the electron beam decreased and the penumbra improved. By irradiating a transparent phantom, the authors also showed that Cherenkov luminescence did not interfere with the detection of air scintillation. In a final illustration of the capabilities of this new technique, the authors visualized air scintillation produced during a total skin irradiation treatment. Air scintillation can be measured to monitor a radiation beam in an inexpensive and nonperturbing manner. This physical phenomenon could be useful for dosimetry of therapeutic radiation beams or for online detection of gross errors during fractionated treatments.

  18. Seeing the invisible: Direct visualization of therapeutic radiation beams using air scintillation

    Energy Technology Data Exchange (ETDEWEB)

    Fahimian, Benjamin; Türkcan, Silvan; Kapp, Daniel S.; Pratx, Guillem, E-mail: pratx@stanford.edu [Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California 94305 (United States); Ceballos, Andrew [Department of Electrical Engineering, Stanford University, Stanford, California 94305 (United States)

    2014-01-15

    Purpose: To assess whether air scintillation produced during standard radiation treatments can be visualized and used to monitor a beam in a nonperturbing manner. Methods: Air scintillation is caused by the excitation of nitrogen gas by ionizing radiation. This weak emission occurs predominantly in the 300–430 nm range. An electron-multiplication charge-coupled device camera, outfitted with an f/0.95 lens, was used to capture air scintillation produced by kilovoltage photon beams and megavoltage electron beams used in radiation therapy. The treatment rooms were prepared to block background light and a short-pass filter was utilized to block light above 440 nm. Results: Air scintillation from an orthovoltage unit (50 kVp, 30 mA) was visualized with a relatively short exposure time (10 s) and showed an inverse falloff (r{sup 2} = 0.89). Electron beams were also imaged. For a fixed exposure time (100 s), air scintillation was proportional to dose rate (r{sup 2} = 0.9998). As energy increased, the divergence of the electron beam decreased and the penumbra improved. By irradiating a transparent phantom, the authors also showed that Cherenkov luminescence did not interfere with the detection of air scintillation. In a final illustration of the capabilities of this new technique, the authors visualized air scintillation produced during a total skin irradiation treatment. Conclusions: Air scintillation can be measured to monitor a radiation beam in an inexpensive and nonperturbing manner. This physical phenomenon could be useful for dosimetry of therapeutic radiation beams or for online detection of gross errors during fractionated treatments.

  19. Homogeneous scintillating LKr/Xe calorimeters

    International Nuclear Information System (INIS)

    Chen, M.; Mullins, M.; Pelly, D.; Shotkin, S.; Sumorok, K.; Akyuz, D.; Chen, E.; Gaudreau, M.P.J.; Bolozdynya, A.; Tchernyshev, V.; Goritchev, P.; Khovansky, V.; Koutchenkov, A.; Kovalenko, A.; Lebedenko, V.; Vinogradov, V.; Gusev, L.; Sheinkman, V.; Krasnokutsky, R.N.; Shuvalov, R.S.; Fedyakin, N.N.; Sushkov, V.; Akopyan, M.; Doke, T.; Kikuchi, J.; Hitachi, A.; Kashiwagi, T.; Masuda, K.; Shibamura, E.; Ishida, N.; Sugimoto, S.

    1993-01-01

    Recent R and D work on full length scintillating homogeneous liquid xenon/krypton (LXe/Kr) cells has established the essential properties for precision EM calorimeters: In-situ calibration using α's, radiation hardness as well as the uniformity required for δE/E≅0.5% for e/γ's above 50 GeV. (orig.)

  20. Fabrication and a Study on the Tapered Scintillator Radiation Sensors

    International Nuclear Information System (INIS)

    Kim, Gye Hong

    2003-02-01

    Optical fibers have been investigated for their uses as sensor materials in various nuclear applications. Comprehensive overviews of their potential usages in nuclear environments can be found in the literatures. Optical fibers with doped scintillating elements in the core have drawn special interests as nuclear radiation detectors. The two important functions of a scintillating optical fiber, as a radiation detector, are scintillator (light emitter) and light transmitter. When radiation interacts with the core material, scintillation occurs and resultant lights are transmitted through the fiber to an opto-electronic device such as a photomultiplier tube. Optical fiber sensors have several advantages as compared to other sensors of conventional material. Since they do not require electric power in the sensor part, they are less susceptible to trouble in harsh environments such as underground or underwater. At relatively low cost a multi-point distributed radiation monitoring system could be made using the fiber sensors. Furthermore, unlike the conventional scintillating counters they are not influenced by any magnetic field surrounding them. This study has been conducted to investigate the feasibility of using scintillating optical fibers for the detection of gamma rays emitted by 137 Cs. Several types of gamma-ray sensors have been constructed by packing different numbers of fibers into aluminum tubes, and tested to detect the 137 Cs gamma ray. During the study it has been found that a tapered fibers might be more efficient to collect the lights produced inside the sensor and transfer them into the transmitting fiber. In order to investigate the effectiveness of the tapered fiber, tapered plastic scintillators, composed of polystyrene with minute amount of dPOPOP and PPO or bPBD, have been fabricated and tested for the detection of gamma rays from 1.0, 1.5, 3.0 and 5.0 μCi 137 Cs sources. The pulse height spectra and the relationship between the radioactivity and

  1. Homogeneous scintillating LKr/Xe calorimeters

    Energy Technology Data Exchange (ETDEWEB)

    Chen, M.; Mullins, M.; Pelly, D.; Shotkin, S.; Sumorok, K. (Lab. for Nuclear Science, MIT, Cambridge, MA (United States)); Akyuz, D.; Chen, E.; Gaudreau, M.P.J. (Plasma Fusion Center, MIT, Cambridge, MA (United States)); Bolozdynya, A.; Tchernyshev, V.; Goritchev, P.; Khovansky, V.; Koutchenkov, A.; Kovalenko, A.; Lebedenko, V.; Vinogradov, V.; Gusev, L.; Sheinkman, V. (ITEP, Moscow (Russia)); Krasnokutsky, R.N.; Shuvalov, R.S.; Fedyakin, N.N.; Sushkov, V. (IHEP, Serpukhov (Russia)); Akopyan, M. (Inst. for Nuclear Research, Moscow (Russia)); Doke, T.; Kikuchi, J.; Hitachi, A.; Kashiwagi, T. (Science and Eng. Res. Lab., Waseda Univ., Tokyo (Japan)); Masuda, K.; Shibamura, E. (Saitama Coll. of Health (Japan)); Ishida, N. (Seikei Univ. (Japan)); Sugimoto, S. (INS, Univ. Tokyo (Japan))

    1993-03-20

    Recent R and D work on full length scintillating homogeneous liquid xenon/krypton (LXe/Kr) cells has established the essential properties for precision EM calorimeters: In-situ calibration using [alpha]'s, radiation hardness as well as the uniformity required for [delta]E/E[approx equal]0.5% for e/[gamma]'s above 50 GeV. (orig.).

  2. Using LEDs to stimulate the recovery of radiation damage to plastic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Wetzel, J., E-mail: james-wetzel@uiowa.edu [The University of Iowa, Iowa City, IA (United States); Tiras, E. [The University of Iowa, Iowa City, IA (United States); Bilki, B. [The University of Iowa, Iowa City, IA (United States); Beykent University, Istanbul (Turkey); Onel, Y. [The University of Iowa, Iowa City, IA (United States); Winn, D. [Fairfield University, Fairfield, CT (United States)

    2017-03-15

    In this study, we consider using LEDs to stimulate the recovery of scintillators damaged from radiation in high radiation environments. We irradiated scintillating tiles of polyethylene naphthalate (PEN), Eljen brand EJ-260 (EJN), an overdoped EJ-260 (EJ2P), and a lab-produced elastomer scintillator (ES) composed of p-terphenyl (ptp) in epoxy. Two different high-dose irradiations took place, with PEN dosed to 100 kGy, and the others to 78 kGy. We found that the ‘blue’ scintillators (PEN and ES) recovered faster and maximally higher with LEDs than without. Conversely exposing the ‘green’ scintillators (EJ-260) to LED light had a nearly negligible effect on the recovery. We hypothesize that the ‘green’ scintillators require wavelengths that match their absorption and emission spectra for LED stimulated recovery.

  3. Quenching the scintillation in CF{sub 4} Cherenkov gas radiator

    Energy Technology Data Exchange (ETDEWEB)

    Blake, T. [Department of Physics, University of Warwick, Coventry (United Kingdom); D' Ambrosio, C. [European Organization for Nuclear Research (CERN), Geneva (Switzerland); Easo, S. [STFC Rutherford Appleton Laboratory, Didcot (United Kingdom); European Organization for Nuclear Research (CERN), Geneva (Switzerland); Eisenhardt, S. [School of Physics and Astronomy, University of Edinburgh, Edinburgh (United Kingdom); Fitzpatrick, C. [Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne (Switzerland); Forty, R.; Frei, C. [European Organization for Nuclear Research (CERN), Geneva (Switzerland); Gibson, V. [Cavendish Laboratory, University of Cambridge, Cambridge (United Kingdom); Gys, T. [European Organization for Nuclear Research (CERN), Geneva (Switzerland); Harnew, N.; Hunt, P. [Department of Physics, University of Oxford, Oxford (United Kingdom); Jones, C.R. [Cavendish Laboratory, University of Cambridge, Cambridge (United Kingdom); Lambert, R.W. [Nikhef National Institute for Subatomic Physics and VU University Amsterdam, Amsterdam (Netherlands); Matteuzzi, C. [Sezione INFN di Milano Bicocca, Milano (Italy); Muheim, F. [School of Physics and Astronomy, University of Edinburgh, Edinburgh (United Kingdom); Papanestis, A., E-mail: antonis.papanestis@stfc.ac.uk [STFC Rutherford Appleton Laboratory, Didcot (United Kingdom); European Organization for Nuclear Research (CERN), Geneva (Switzerland); Perego, D.L. [Sezione INFN di Milano Bicocca, Milano (Italy); Università di Milano Bicocca, Milano (Italy); Piedigrossi, D. [European Organization for Nuclear Research (CERN), Geneva (Switzerland); Plackett, R. [Imperial College London, London (United Kingdom); Powell, A. [Department of Physics, University of Oxford, Oxford (United Kingdom); and others

    2015-08-11

    CF{sub 4} is used as a Cherenkov gas radiator in one of the Ring Imaging Cherenkov detectors at the LHCb experiment at the CERN Large Hadron Collider. CF{sub 4} is well known to have a high scintillation photon yield in the near and far VUV, UV and in the visible wavelength range. A large flux of scintillation photons in our photon detection acceptance between 200 and 800 nm could compromise the particle identification efficiency. We will show that this scintillation photon emission system can be effectively quenched, consistent with radiationless transitions, with no significant impact on the photons resulting from Cherenkov radiation.

  4. Test of a Fiber Optic-Based LYSO Scintillator Dosimeter in a 60Co Irradiation Chamber

    International Nuclear Information System (INIS)

    Kim, Tae Hyoung; Kim, Jae Kyung; Park, Jae Woo

    2010-01-01

    Due to its excellent remote measurability and high spatial resolution, the fiber optic-based radiation dosimeter has been extensively explored for its usability in medical applications by several researchers. In the previous work, we reported the result of our investigation on feasibility of a photon dosimeter constructed with a BGO(Bi 4 Ge 3 O 12 ) or GSO(Gd 2 SiO 5 ) scintillator piece coupled to a plastic optical fiber. The plastic optical fiber had a diameter of 3mm and the scintillator piece was in a cylindrical form with 5mm diameter. The size of the scitillator piece as well as the fiber should be as small as possible for higher spatial resolution, and the radiation hardness should be high enough for stable operation in strong radiation fields. Recently, LYSO(Cerium-doped Lutetium Yttrium Orthosilicate) scintillators, which have much higher light yield and radiation hardness than BGO and GSO, have been commercially available. This paper reports the result of our investigation on dosimetric characteristics of a fiber optic-based dosimeter employing a smaller LYSO scintillator piece with 2mm diameter coupled to a silica optical fiber with 1mm core diameter

  5. Continuing Studies on Lead/Scintillating Fibres Calorimetry (LFC)

    CERN Multimedia

    2002-01-01

    Starting from the results obtained in the framework of the LAA Project~2B, we propose a continuation of the R&D on lead/scintillating fibres calorimetry (``spaghetti calorimetry''), including further tests on the old calorimeter prototypes and the construction and testing of new prototypes. The main results we pursue concern the performances of a projective calorimeter built with new, cheaper, techniques and the radiation hardness of the scintillating fibres, the optimization of a preshower detector system is also studied.

  6. Development of a hard x-ray focal plane compton polarimeter: a compact polarimetric configuration with scintillators and Si photomultipliers

    Science.gov (United States)

    Chattopadhyay, T.; Vadawale, S. V.; Goyal, S. K.; Mithun, N. P. S.; Patel, A. R.; Shukla, R.; Ladiya, T.; Shanmugam, M.; Patel, V. R.; Ubale, G. P.

    2016-02-01

    X-ray polarization measurement of cosmic sources provides two unique parameters namely degree and angle of polarization which can probe the emission mechanism and geometry at close vicinity of the compact objects. Specifically, the hard X-ray polarimetry is more rewarding because the sources are expected to be intrinsically highly polarized at higher energies. With the successful implementation of Hard X-ray optics in NuSTAR, it is now feasible to conceive Compton polarimeters as focal plane detectors. Such a configuration is likely to provide sensitive polarization measurements in hard X-rays with a broad energy band. We are developing a focal plane hard X-ray Compton polarimeter consisting of a plastic scintillator as active scatterer surrounded by a cylindrical array of CsI(Tl) scintillators. The scatterer is 5 mm diameter and 100 mm long plastic scintillator (BC404) viewed by normal PMT. The photons scattered by the plastic scatterer are collected by a cylindrical array of 16 CsI(Tl) scintillators (5 mm × 5 mm × 150 mm) which are read by Si Photomultiplier (SiPM). Use of the new generation SiPMs ensures the compactness of the instrument which is essential for the design of focal plane detectors. The expected sensitivity of such polarimetric configuration and complete characterization of the plastic scatterer, specially at lower energies have been discussed in [11, 13]. In this paper, we characterize the CsI(Tl) absorbers coupled to SiPM. We also present the experimental results from the fully assembled configuration of the Compton polarimeter.

  7. Optimization of the light extraction from heavy inorganic scintillators

    CERN Document Server

    Kronberger, Matthias; Lecoq, P

    2008-01-01

    Inorganic scintillators are widely used in modern medical imaging modalities as converter for the X- and gamma-radiation that is used to obtain information about the interior of the body. Likewise, they are applied in high-energy physics to measure the energy of particles that are produced in particle physics experiments. Their use is motivated by the very good detection efficiency of these materials for hard radiation which allows the construction of relatively compact and finely pixelised systems with a high spatial resolution. One key problem in the development of the next generation of particle detectors and medical imaging systems is the optimisation of the energy resolution of the detectors. This parameter is influenced by the statistical fluctuations of the light output of the scintillators, i.e. by the number of photons that are detected when a particle deposits its energy in the scintillator. The light output of the scintillator depends not only on the absolute number of generated photons but also on...

  8. Time of gamma radiation transport in NaI(Tl) scintillators

    International Nuclear Information System (INIS)

    Matarrita, A.S.

    1976-01-01

    The distribution of time intervals in function of the energy, ocurred between the incidence of a gamma radiation in the face of a scintillation crystal and the arrival of the correspondent scintillation at photocathode, is calculated. The mean square fluctuation these time interval, relating to resolution of the detector system is determined. The calculations are done for NaI (Tl) cylindrical crystals with the radiation source placed in the symmetry axis, in two situations: para axial incidence and oblique incidence, indicating a good agreement with experimental data. (M.C.K.) [pt

  9. Radiation-Hard Quartz Cerenkov Calorimeters

    International Nuclear Information System (INIS)

    Akgun, U.; Onel, Y.

    2006-01-01

    New generation hadron colliders are going to reach unprecedented energies and radiation levels. Quartz has been identified as a radiation-hard material that can be used for Cerenkov calorimeters of the future experiments. We report from the radiation hardness tests performed on quartz fibers, as well as the characteristics of the quartz fiber and plate Cerenkov calorimeters that have been built, designed, and proposed for the CMS experiment

  10. CMOS optimization for radiation hardness

    International Nuclear Information System (INIS)

    Derbenwick, G.F.; Fossum, J.G.

    1975-01-01

    Several approaches to the attainment of radiation-hardened MOS circuits have been investigated in the last few years. These have included implanting the SiO 2 gate insulator with aluminum, using chrome-aluminum layered gate metallization, using Al 2 O 3 as the gate insulator, and optimizing the MOS fabrication process. Earlier process optimization studies were restricted primarily to p-channel devices operating with negative gate biases. Since knowledge of the hardness dependence upon processing and design parameters is essential in producing hardened integrated circuits, a comprehensive investigation of the effects of both process and design optimization on radiation-hardened CMOS integrated circuits was undertaken. The goals are to define and establish a radiation-hardened processing sequence for CMOS integrated circuits and to formulate quantitative relationships between process and design parameters and the radiation hardness. Using these equations, the basic CMOS design can then be optimized for radiation hardness and some understanding of the basic physics responsible for the radiation damage can be gained. Results are presented

  11. Radiation-resistant composite scintillators based on GSO and GPS grains

    Energy Technology Data Exchange (ETDEWEB)

    Boyarintsev, A.Yu. [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine); Galunov, N.Z. [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine); V.N. Karasin Kharkov National University, 4 Svobody Sq., 61022 Kharkiv (Ukraine); Gerasymov, Ia.V.; Karavaeva, N.L. [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine); Krech, A.V., E-mail: AntonKrech@gmail.com [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine); Levchuk, L.G.; Popov, V.F. [National Science Center, Kharkov Institute of Physics and Technology, 1 Akademicheskaya Str., 61108 Kharkiv (Ukraine); Sidletskiy, O.Ts. [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine); Sorokin, P.V. [National Science Center, Kharkov Institute of Physics and Technology, 1 Akademicheskaya Str., 61108 Kharkiv (Ukraine); Tarasenko, O.A. [Institute for Scintillation Materials, National Academy of Sciences of Ukraine, 60 Nauki Avenue, 61001 Kharkiv (Ukraine)

    2017-01-01

    The effect of irradiation on the scintillation light output, optical transmittance, and luminescent spectra of composite scintillators based on grains of single crystals Gd{sub 2}SiO{sub 5}:Ce (GSO) and Gd{sub 2}Si{sub 2}O{sub 7}:Ce (GPS) is studied. The dielectric gel Sylgard-184 is the base and the binder for the grains inside the composite scintillator. The paper presents and analyzes the results obtained for the scintillators exposed by 10 MeV electrons from the linear electron accelerator at room temperature. The exposure doses D≤250 Mrad. The dose rate is 0.2 or 1500 Mrad/h. The study has shown that the composite scintillators based on the grains of GSO and GPS are radiation-resistant over the range of the irradiation.

  12. Evaluation of the Detection Efficiency of LYSO Scintillator in the Fiber-Optic Radiation Sensor

    Directory of Open Access Journals (Sweden)

    Chan Hee Park

    2014-01-01

    Full Text Available The aim of this study was to develop and evaluate fiber-optic sensors for the remote detection of gamma rays in areas that are difficult to access, such as a spent fuel pool. The fiber-optic sensor consists of a light-generating probe, such as scintillators for radiation detection, plastic optical fibers, and light-measuring devices, such as PMT. The (Lu,Y2SiO5:Ce(LYSO:Ce scintillator was chosen as the light-generating probe. The (Lu,Y2SiO5:Ce(LYSO:Ce scintillator has higher scintillation efficiency than the others and transmits light well through an optical fiber because its refraction index is similar to the refractive index of the optical fiber. The fiber-optic radiation sensor using the (Lu,Y2SiO5:Ce(LYSO:Ce scintillator was evaluated in terms of the detection efficiency and reproducibility for examining its applicability as a radiation sensor.

  13. Ultrahigh resolution radiation imaging system using an optical fiber structure scintillator plate.

    Science.gov (United States)

    Yamamoto, Seiichi; Kamada, Kei; Yoshikawa, Akira

    2018-02-16

    High resolution imaging of radiation is required for such radioisotope distribution measurements as alpha particle detection in nuclear facilities or high energy physics experiments. For this purpose, we developed an ultrahigh resolution radiation imaging system using an optical fiber structure scintillator plate. We used a ~1-μm diameter fiber structured GdAlO 3 :Ce (GAP) /α-Al 2 O 3 scintillator plate to reduce the light spread. The fiber structured scintillator plate was optically coupled to a tapered optical fiber plate to magnify the image and combined with a lens-based high sensitivity CCD camera. We observed the images of alpha particles with a spatial resolution of ~25 μm. For the beta particles, the images had various shapes, and the trajectories of the electrons were clearly observed in the images. For the gamma photons, the images also had various shapes, and the trajectories of the secondary electrons were observed in some of the images. These results show that combining an optical fiber structure scintillator plate with a tapered optical fiber plate and a high sensitivity CCD camera achieved ultrahigh resolution and is a promising method to observe the images of the interactions of radiation in a scintillator.

  14. Semiconductor scintillator detector for gamma radiation

    International Nuclear Information System (INIS)

    Laan, F.T.V. der; Borges, V.; Zabadal, J.R.S.

    2015-01-01

    Nowadays the devices employed to evaluate individual radiation exposition are based on dosimetric films and thermoluminescent crystals, whose measurements must be processed in specific transductors. Hence, these devices carry out indirect measurements. Although a new generation of detectors based on semiconductors which are employed in EPD's (Electronic Personal Dosemeters) being yet available, it high producing costs and large dimensions prevents the application in personal dosimetry. Recent research works reports the development of new detection devices based on photovoltaic PIN diodes, which were successfully employed for detecting and monitoring exposition to X rays. In this work, we step forward by coupling a 2mm anthracene scintillator NE1, which converts the high energy radiation in visible light, generating a Strong signal which allows dispensing the use of photomultipliers. A low gain high performance amplifier and a digital acquisition device are employed to measure instantaneous and cumulative doses for energies ranging from X rays to Gamma radiation up to 2 MeV. One of the most important features of the PIN diode relies in the fact that it can be employed as a detector for ionization radiation, since it requires a small energy amount for releasing electrons. Since the photodiode does not amplify the corresponding photon current, it must be coupled to a low gain amplifier. Therefore, the new sensor works as a scintillator coupled with a photodiode PIN. Preliminary experiments are being performed with this sensor, showing good results for a wide range of energy spectrum. (author)

  15. Luminescence and energy transfer in Garnet Scintillators

    OpenAIRE

    Ogiegło, J.M.

    2012-01-01

    The thesis is focused on development and fundamental understanding of scintillators that play a central role in the field of medical imaging. These materials convert high energy, gamma or X-ray, radiation into visible light that is then used to create a detailed image of the patient’s body. The power of such imaging techniques as diagnostic medical tools is hard to overestimate.

  16. Radiation damage studies on the optical and mechanical properties of plastic scintillators

    International Nuclear Information System (INIS)

    Mizue Hamada, Margarida; Roberto Rela, Paulo; Eduardo da Costa, Fabio; Henrique de Mesquita, Carlos

    1999-01-01

    This paper describes the radiation damage studies on a large volume plastic scintillator based in polystyrene doped with PPO and POPOP. The consequences on their mechanical and scintillation properties were evaluated before and after irradiation with different dose rates of 60 Co gamma radiation, in several doses. The optical results show a significant difference in the radiation susceptibility, when the plastic scintillator is irradiated at low rate (0.1 kGy/h) with that irradiated at high dose rate (85 kGy/h). The losses in the optical and mechanical properties increase as the irradiation dose is increased. The damage evaluated by the transmittance, emission intensity, pulse height and tensile strength was normalized as a damage fraction and fitted by a bi-exponential function. It was observed that the damage for irradiation is not permanent and it obeys a bi-exponential function

  17. Design of new polymers to improve radiation stability of plastic scintillators

    International Nuclear Information System (INIS)

    Barashkov, N.; Korotkikh, N.; Novikova, T.; Markley, F.; Pla-Dalmau, A.; Foster, G.; Rivard, M.

    1993-11-01

    Diffusion enhancers have been used to increase the radiation stability of scintillators. There is a danger that such additions will also allow the scintillation dyes to diffuse out of the matrix. Covalent bonding of the dyes into the matrix would eliminate any luminophore migration. We have demonstrated that dyes with vinyl and divinyl groups can be successfully bonded into a styrene matrix. Both primary and secondary luminophores have been bonded into the matrix and in several cases they have been compared with similar luminophores without the vinyl groups in otherwise identical compositions. In general the bonding does not greatly affect either the radiation stability of the composition or its light output, but is therefore proven to be an acceptable method of preventing luminophore migration. Scintillators with a base of aromatic or alicyclic epoxy resins have also been made with bonded luminophores. In these compositions the bonding is accomplished by using luminophores with one or two amine groups or epoxy groups. When using the amines, they are first prereacted with the epoxy resin and the modified epoxy is then cured with hexahydrophthalic anhydride or trimethoxyboroxine. The emission and absorption spectra of many of these reactive luminophores are given, along with the light outputs and radiation stability of the resulting scintilators

  18. Design of new polymers to improve radiation stability of plastic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Barashkov, N.; Korotkikh, N.; Novikova, T. [Karpov Inst. of Physical Chemistry (Russian Federation); Markley, F.; Pla-Dalmau, A.; Foster, G. [Fermi National Accelerator Lab., Batavia, IL (United States); Rivard, M. [Michigan Univ., Ann Arbor, MI (United States)

    1993-11-01

    Diffusion enhancers have been used to increase the radiation stability of scintillators. There is a danger that such additions will also allow the scintillation dyes to diffuse out of the matrix. Covalent bonding of the dyes into the matrix would eliminate any luminophore migration. We have demonstrated that dyes with vinyl and divinyl groups can be successfully bonded into a styrene matrix. Both primary and secondary luminophores have been bonded into the matrix and in several cases they have been compared with similar luminophores without the vinyl groups in otherwise identical compositions. In general the bonding does not greatly affect either the radiation stability of the composition or its light output, but is therefore proven to be an acceptable method of preventing luminophore migration. Scintillators with a base of aromatic or alicyclic epoxy resins have also been made with bonded luminophores. In these compositions the bonding is accomplished by using luminophores with one or two amine groups or epoxy groups. When using the amines, they are first prereacted with the epoxy resin and the modified epoxy is then cured with hexahydrophthalic anhydride or trimethoxyboroxine. The emission and absorption spectra of many of these reactive luminophores are given, along with the light outputs and radiation stability of the resulting scintilators.

  19. Response of CsI:Pb Scintillator Crystal to Neutron Radiation

    Science.gov (United States)

    Costa Pereira, Maria da Conceição; Filho, Tufic Madi; Berretta, José Roberto; Náhuel Cárdenas, José Patrício; Iglesias Rodrigues, Antonio Carlos

    2018-01-01

    The helium-3 world crisis requires a development of new methods of neutron detection to replace commonly used 3He proportional counters. In the past decades, great effort was made to developed efficient and fast scintillators to detect radiation. The inorganic scintillator may be an alternative. Inorganic scintillators with much higher density should be selected for optimal neutron detection efficiency taking into consideration the relevant reactions leading to light emission. These detectors should, then, be carefully characterized both experimentally and by means of advanced simulation code. Ideally, the detector should have the capability to separate neutron and gamma induced events either by amplitude or through pulse shape differences. As neutron sources also generate gamma radiation, which can interfere with the measurement, it is necessary that the detector be able to discriminate the presence of such radiation. Considerable progress has been achieved to develop new inorganic scintillators, in particular increasing the light output and decreasing the decay time by optimized doping. Crystals may be found to suit neutron detection. In this report, we will present the results of the study of lead doped cesium iodide crystals (CsI:Pb) grown in our laboratory, using the vertical Bridgman technique. The concentration of the lead doping element (Pb) was studied in the range 5x10-4 M to 10-2 M . The crystals grown were subjected to annealing (heat treatment). In this procedure, vacuum of 10-6 mbar and continuous temperature of 350°C, for 24 hours, were employed. In response to neutron radiation, an AmBe source with energy range of 1 MeV to 12 MeV was used. The activity of the AmBe source was 1Ci Am. The fluency was 2.6 x 106 neutrons/second. The operating voltage of the photomultiplier tube was 1700 V; the accumulation time in the counting process was 600 s and 1800 s. The scintillator crystals used were cut with dimensions of 20 mm diameter and 10 mm height.

  20. Scintillation characteristics of phosphich-detector for detection of beta- and gamma-radiations

    CERN Document Server

    Ananenko, A A; Gavrilyuk, V

    2002-01-01

    The results of the study on the influence of individual peculiarities of the compound scintillation detector structure on the value and stability of the light yield by the gamma- and beta-radiation combined registration are presented. The phosphich detector is manufactured from the sodium iodide monocrystal, activated by thallium, and the scintillation plastic on the polystyrol basis. The comparison of the experimental results with the mathematical modeling data revealed certain regularities of the process of forming the phosphich detector light signal. The recommendations are worked out by means whereof the following characteristics of the scintillation unit: the light yield and its stability, amplitude resolution and the peak-to-valley ratio by the gamma- and beta-radiation registration were improved

  1. Heat- and radiation-resistant scintillator for electron microscopes

    International Nuclear Information System (INIS)

    Kosov, A.V.; Petrov, S.A.; Puzyr', A.P.; Chetvergov, N.A.

    1987-01-01

    The use of a scintillator consisting of a single crystal of bismuth orthogermanate, which has high heat and radiation resistance, in REM-100, REM-200, and REM-100U electron microscopes is described. A study of the heat and radiation stabilities of single crystals of bismuth orthogermanate (Bi 4 Ge 3 O 12 ) has shown that they withstood multiple electron-beam heating redness (T ∼ 800 0 C) without changes in their properties

  2. Current trends in scintillator detectors and materials

    International Nuclear Information System (INIS)

    Moses, W.W.

    2002-01-01

    The last decade has seen a renaissance in inorganic scintillator development for gamma ray detection. Lead tungstate (PbWO 4 ) has been developed for high-energy physics experiments, and possesses exceptionally high density and radiation hardness, albeit with low luminous efficiency. Lutetium orthosilicate or LSO (Lu 2 SiO 5 :Ce) possesses a unique combination of high luminous efficiency, high density, and reasonably short decay time, and is now incorporated in commercial positron emission tomography cameras. There have been advances in understanding the fundamental mechanisms that limit energy resolution, and several recently discovered materials (such as LaBr 3 :Ce) possess energy resolution that approaches that of direct solid state detectors. Finally, there are indications that a neglected class of scintillator materials that exhibit near band-edge fluorescence could provide scintillators with sub-nanosecond decay times and high luminescent efficiency

  3. Scintillation counting apparatus

    International Nuclear Information System (INIS)

    Noakes, J.E.

    1978-01-01

    Apparatus is described for the accurate measurement of radiation by means of scintillation counters and in particular for the liquid scintillation counting of both soft beta radiation and gamma radiation. Full constructional and operating details are given. (UK)

  4. Radiation detectors based by polymer materials

    International Nuclear Information System (INIS)

    Cherestes, Margareta; Cherestes, Codrut; Constantinescu, Livia

    2004-01-01

    Scintillation counters make use of the property of certain chemical compounds to emit short light pulses after excitation produced by the passage of charged particles or photons of high energy. These flashes of light are detected by a photomultiplier tube that converts the photons into a voltage pulse. The light emitted from the detector also can be collected, focussed and dispersed by a CCD detector. The study of the evolution of the light emission and of the radiation damage under irradiation is a primary topic in the development of radiation hard polymer based scintillator. Polymer scintillator thin films are used in monitoring radiation beam intensities and simultaneous counting of different radiations. Radiation detectors have characteristics which depend on: the type of radiation, the energy of radiation, and the material of the detector. Three types of polymer thin films were studied: a polyvinyltoluene based scintillator, fluorinated polyimide and PMMA. (authors)

  5. Radiation damage of tile/fiber scintillator modules for the SDC calorimeter

    International Nuclear Information System (INIS)

    Hu, L.; Liu, N.; Mao, H.; Tan, Y.; Wang, G.; Zhang, C.; Zhang, G.; Zhang, L.; Zhang, Z.; Zhao, X.; Zheng, L.; Zhong, X.; Zhou, Y.; Han, S.; Byon, A.; Green, D.; Para, A.; Johnson, K.; Barnes, V.

    1992-02-01

    The measurements of radiation damage of tile/fiber scintillator modules to be used for the SDC calorimeter are described. Four tile/fiber scintillator modules were irradiated up to 6 Mrad with the BEPC 1.1 GeV electron beam. We have studied the light output at different depths in the modules and at different integrated doses, the recovery process and the dependence on the ambient atmosphere

  6. Highly lead-loaded red plastic scintillators as an X-ray imaging system for the laser Mega Joule

    International Nuclear Information System (INIS)

    Hamel, Matthieu; Normand, Stephane; Turk, Gregory; Darbon, Stephane

    2012-01-01

    The scope of this project intends to record spatially resolved images of core shape and size of a deuterium-tritium micro-balloon during inertial confinement fusion (ICF) experiments at Laser Mega Joule facility (LMJ). We need to develop an x-ray imaging system which can operate in the hard radiative background generated by an ignition shot of ICF. The scintillator is a part of the imaging system and has to gather a compromise of scintillating properties (scintillating efficiency, decay time, emission wavelength) so as to both operate in the hard radiative environment and to allow the acquisition of spatially resolved images. Inorganic scintillators cannot be used because no compromise can be found regarding the expected scintillating properties. Most of them are not fast enough and emit blue light. Organic scintillators are generally fast, but present low x-ray photoelectric absorption in the 10 to 40 keV range. This does not enable the acquisition of spatially resolved images. To this aim, we have developed highly lead-loaded and red-fluorescent fast plastic scintillators. Such a combination is not currently available via scintillator suppliers, since they propose only blue-fluorescent plastic scintillators doped with up to 12 wt% Pb. Thus, incorporation ratio up to 27 wt% Pb has been reached in our laboratory, which can afford a plastic scintillator with an outstanding Z(eff) close to 50. X-rays in the 10 to 40 keV range can thus interact with a higher probability of photoelectric effect than for classic organic scintillators, such as NE-102. The strong orange-red fluorescence can be filtered, so that we can eliminate residual Cerenkov light, generated by gamma-ray absorption in glass parts of the imaging system. Characteristic decay times of our scintillators evaluated under UV excitation were estimated to be in the range 10 to 13 ns. (authors)

  7. Scintillator device using a combined organic-inorganic scintillator as dose ratemeter

    International Nuclear Information System (INIS)

    Kolb, W.; Lauterbach, U.

    1974-01-01

    The dose ratemeter independent of energy in the energy region 17 keV to 3 MeV consists of an organic and an inorganic scintillator. The organic scintillation material of an anthracene monocrystal is surrounded by ZnS surface coating. The coating thickness of the inorganic scintillator ZnS is measured in such a manner for gamma and X-radiation below 100 keV that the light produced due to the incident radiation compensates for the decrease of light produced in the organic scintillator. The whole energy and dose rate region of interest for radiation protection can thus be measured with a detector volume of 135 cm 3 . (DG) [de

  8. Radiation distribution measurement using plastic scintillating optical fibers for survey of radioactive contamination in wide area

    International Nuclear Information System (INIS)

    Ito, Chikara; Ito, Keisuke; Ishikawa, Takashi; Yoshida, Akihiro; Sanada, Yukihisa; Torii, Tatsuo; Nohtomi, Akihiro; Wakabayashi, Genichiro; Miyazaki, Nobuyuki

    2013-01-01

    It is important to examine distribution of environmental contamination due to the accident of Fukushima Daiichi Nuclear Power Station and to confirm the effect of decontamination works. We have applied radiation distribution measurement using plastic scintillating optical fibers (PSFs) in the survey of contamination in wide area including residential, farmland, forests, etc. In the measurements system, two scintillation lights that emitted at an incidence of a radiation transmit to photomultiplier tubes at the both end of PSFs. The position where scintillation light emitted is obtained from the detection time difference of each photomultiplier tube. The distribution of light emission quantity indicates the distribution of radiation incident in a PSF which is corresponds to the distribution of dose-rate. The radiation detection system using the PSFs has been applied to the radiation distribution measurement on grounds, trees, etc. The results show a good agreement with point data measured by survey meters using sodium iodide scintillators. As the PSFs which have water resistance, they have been successfully applied to the radiation distribution measurement in the river. We have also succeeded in measuring two-dimensional distribution of radiation by measuring the count rate while moving to the fiber at a constant speed. (author)

  9. Ticor-based scintillation detectors for detection of mixed radiation

    CERN Document Server

    Litvinov, L A; Kolner, V B; Ryzhikov, V D; Volkov, V G; Tarasov, V A; Zelenskaya, O V

    2002-01-01

    Detection of mixed radiation of thermal neutrons and gamma-rays have been realized using a new ceramic material based on small-crystalline long-wave scintillator alpha-Al sub 2 O sub 3 :Ti (Ticor) and lithium fluoride. Characteristics are presented for scintillators with Si-PIN-PD type photoreceivers and PMT under sup 2 sup 3 sup 9 Pu alpha-particles, sup 2 sup 0 sup 7 Bi internal conversion electrons,as well as sup 2 sup 4 sup 1 Am and sup 1 sup 3 sup 7 Cs gamma-quanta. Detection efficiency of thermal neutron is estimated for composite materials based on Ticor and lithium fluoride.

  10. Impact of aging on radiation hardness

    International Nuclear Information System (INIS)

    Shaneyfelt, M.R.; Winokur, P.S.; Fleetwood, D.M.

    1997-01-01

    Burn-in effects are used to demonstrate the potential impact of thermally activated aging effects on functional and parametric radiation hardness. These results have implications on hardness assurance testing. Techniques for characterizing aging effects are proposed

  11. Optimization of the scintillation parameters of the lead tungstate crystals for their application in high precision electromagnetic calorimetry; Optimisation des parametres de scintillation des cristaux de tungstate de plomb pour leur application dans la calorimetrie electromagnetique de haute precision

    Energy Technology Data Exchange (ETDEWEB)

    Drobychev, G

    2000-04-12

    In the frame of this dissertation work scintillation properties of the lead tungstate crystals (PWO) and possibilities of their use were studied foreseeing their application for electromagnetic calorimetry in extreme radiation environment conditions of new colliders. The results of this work can be summarized in the following way. 1. A model of the scintillations origin in the lead tungstate crystals which includes processes influencing on the crystals radiation hardness and presence of slow components in scintillations was developed. 2. An analysis of the influences of the PWO scintillation properties changes on the parameters of the electromagnetic calorimeter was done. 3. Methods of the light collection from the large scintillation elements of complex shape made of the birefringent scintillation crystal with high refraction index and low light yield in case of signal registration by a photodetector with sensitive surface small in compare with the output face of scintillator were Studied. 4. Physical principles of the methodology of the scintillation crystals certification during their mass production foreseeing their installation into a calorimeter electromagnetic were developed. Correlations between the results of measurements of the PWO crystals parameters by different methods were found. (author)

  12. An ideal scintillator – ZnO:Sc for sub-nanosecond pulsed radiation detection

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, Kan, E-mail: zhangkan8414@163.com [Northwest Institute of Nuclear Technology, Xi’an 710024 (China); Ouyang, Xiaoping [Northwest Institute of Nuclear Technology, Xi’an 710024 (China); Xi’an Jiaotong University, Xi’an 710049 (China); Song, Zhaohui; Han, Hetong [Northwest Institute of Nuclear Technology, Xi’an 710024 (China); Zuo, Yanbin [China Nonferrous Metal Guilin Research Institute of Geology for Mineral Resource, Guilin 541004 (China); Guan, Xingyin [Northwest Institute of Nuclear Technology, Xi’an 710024 (China); Xi’an Jiaotong University, Xi’an 710049 (China); Tan, Xinjian; Zhang, Zichuan; Liu, Junhong [Northwest Institute of Nuclear Technology, Xi’an 710024 (China)

    2014-08-21

    ZnO-based scintillators are particularly well suited for use as ultrafast pulsed radiation detectors which have shown broad application prospects in various fields such as the inertial confinement fusion (ICF) diagnosis, the nuclear reaction mechanism, etc. Using the hydro-thermal method, a ZnO single-crystal doped with Scandium (ZnO:Sc) sample was prepared. As a new ZnO-based scintillator, the scintillation characteristics of ZnO:Sc have not been reported previously. In this paper, optical and scintillation characteristics of ZnO:Sc single-crystal were studied. Also a scintillation detector based on ZnO:Sc was designed. Excited by the alpha-particle, the rise time of ZnO:Sc detectors was from 162.5 to 170.7 ps, and the fall time was from 300.4 to 328.8 ps.

  13. An ideal scintillator – ZnO:Sc for sub-nanosecond pulsed radiation detection

    International Nuclear Information System (INIS)

    Zhang, Kan; Ouyang, Xiaoping; Song, Zhaohui; Han, Hetong; Zuo, Yanbin; Guan, Xingyin; Tan, Xinjian; Zhang, Zichuan; Liu, Junhong

    2014-01-01

    ZnO-based scintillators are particularly well suited for use as ultrafast pulsed radiation detectors which have shown broad application prospects in various fields such as the inertial confinement fusion (ICF) diagnosis, the nuclear reaction mechanism, etc. Using the hydro-thermal method, a ZnO single-crystal doped with Scandium (ZnO:Sc) sample was prepared. As a new ZnO-based scintillator, the scintillation characteristics of ZnO:Sc have not been reported previously. In this paper, optical and scintillation characteristics of ZnO:Sc single-crystal were studied. Also a scintillation detector based on ZnO:Sc was designed. Excited by the alpha-particle, the rise time of ZnO:Sc detectors was from 162.5 to 170.7 ps, and the fall time was from 300.4 to 328.8 ps

  14. Development of radiation hard components for remote maintenance

    International Nuclear Information System (INIS)

    Oka, Kiyoshi; Obara, Kenjiro; Kakudate, Satoshi; Tominaga, Ryuichiro; Akada, Tamio; Morita, Hirosuke.

    1997-01-01

    In International Thermonuclear Experimental Reactor (ITER), in-vessel remote-handling is inevitably required to assemble and maintain activated in-vessel components due to D-T operation. The components of the in-vessel remote-handling system must have sufficient radiation hardness to allow for operation under an intense gamma-ray radiation of over 30 kGy/h for periods up to more than 1,000 hours. To this end, extensive irradiation tests and quality improvements including the optimization of material composition have been conducted through the ITER R and D program in order to develop radiation hard components which satisfy radiation doses from 10 MGy to 100 MGy at the dose rate of 10 kGy/h. This paper outlines the latest status of the radiation hard component development that has been conducted as the Japan Home Team's contribution to ITER. The remote-handling components tested are categorized into either robotics, viewing systems or common components. The irradiation tests include commercial base products for screening both modified and newly developed products to improve their radiation hardness. (author)

  15. Development of radiation hardness components for ITER remote maintenance

    Energy Technology Data Exchange (ETDEWEB)

    Obara, Kenjiro; Kakudate, Satoshi; Oka, Kiyoshi; Ito, Akira [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment; Yagi, Toshiaki; Morita, Yousuke

    1998-04-01

    In the ITER, in-vessel remote handling is required to assemble and maintain in-vessel components in DT operations. Since in-vessel remote handling systems must operate under intense gamma ray radiation exceeding 30 kGy/h, their components must have sufficiently high radiation hardness to allow maintenance long enough in ITER in-vessel environments. Thus, extensive radiation tests and quality improvement, including optimization of material compositions, have been conducted through the ITER R and D program to develop radiation hardness components that meet radiation doses from 10 to 100 MGy at 10 kGy/h. This paper presents the latest on radiation hardness component development conducted by the Japan Home Team as a contribution to the ITER. The remote handling components tested are categorized for use in robotic or viewing systems, or as common components. Radiation tests have been conducted on commercially available products for screening, on modified products, and on new products to improve the radiation hardness. (author)

  16. Development of radiation hardness components for ITER remote maintenance

    International Nuclear Information System (INIS)

    Obara, Kenjiro; Kakudate, Satoshi; Oka, Kiyoshi; Ito, Akira; Yagi, Toshiaki; Morita, Yousuke

    1998-01-01

    In the ITER, in-vessel remote handling is required to assemble and maintain in-vessel components in DT operations. Since in-vessel remote handling systems must operate under intense gamma ray radiation exceeding 30 kGy/h, their components must have sufficiently high radiation hardness to allow maintenance long enough in ITER in-vessel environments. Thus, extensive radiation tests and quality improvement, including optimization of material compositions, have been conducted through the ITER R and D program to develop radiation hardness components that meet radiation doses from 10 to 100 MGy at 10 kGy/h. This paper presents the latest on radiation hardness component development conducted by the Japan Home Team as a contribution to the ITER. The remote handling components tested are categorized for use in robotic or viewing systems, or as common components. Radiation tests have been conducted on commercially available products for screening, on modified products, and on new products to improve the radiation hardness. (author)

  17. Scintillation camera for establishing the coordinates of a radiation stimuli produced by a radiation field

    International Nuclear Information System (INIS)

    Zioni, J.; Klein, Y.; Inbar, D.

    1977-01-01

    A scintillation camera has a planar scintillating crystal that produces light events whose spatial distribution corresponds to the spatial distribution of the radiation stimuli causing such events, and a plurality of photomultipliers having photocathodes for receiving light from the crystal through a planar face thereof. Computing circuitry coupled to the photomultipliers computes the projection of a light event in the crystal on a reference axis by forming an analytical function of the outputs of the photomultipliers according to the spatial location of the light event in the crystal

  18. Investigating the effect of K-characteristic radiation on the performance of nuclear medicine scintillators by Monte Carlo methods

    International Nuclear Information System (INIS)

    Liaparinos, Panagiotis; Kandarakis, Ioannis; Cavouras, Dionisis; Delis, Harry; Panayiotakis, George

    2006-01-01

    The aim of this study was to evaluate the effect of K-characteristic radiation on the performance of scintillator crystals incorporated in nuclear medicine detectors (LSO, BGO, GSO). K-characteristic radiation is produced within materials of at least one high atomic number element (e.g. Lu, Gd, Bi). This radiation may either be reabsorbed or it may escape the scintillator. In both cases the light emission efficiency of the scintillator may be affected resulting in either spatial or energy resolution degradation. A computational program, based on Monte Carlo methods, was developed in order to simulate the transport of K-characteristic radiation within the most commonly used scintillator materials. Crystal thickness was allowed to vary from 0.5 up to 15 mm. A monoenergetic pencil beam, with energy varying from 0.60 to 0.511 MeV was considered to fall on the center of the crystal surface. The dominant γ-ray interactions (elastic and inelastic scattering and photoelectric absorption) were taken into account in the simulation. Results showed that, depending on crystal thickness, incident photon energy and scintillator's intrinsic properties (L or K-fluorescence yield, effective atomic number and density), the scintillator's emission efficiency may be significantly reduced and affect spatial or energy resolution

  19. Plastic scintillation dosimetry: Optimal selection of scintillating fibers and scintillators

    International Nuclear Information System (INIS)

    Archambault, Louis; Arsenault, Jean; Gingras, Luc; Sam Beddar, A.; Roy, Rene; Beaulieu, Luc

    2005-01-01

    Scintillation dosimetry is a promising avenue for evaluating dose patterns delivered by intensity-modulated radiation therapy plans or for the small fields involved in stereotactic radiosurgery. However, the increase in signal has been the goal for many authors. In this paper, a comparison is made between plastic scintillating fibers and plastic scintillator. The collection of scintillation light was measured experimentally for four commercial models of scintillating fibers (BCF-12, BCF-60, SCSF-78, SCSF-3HF) and two models of plastic scintillators (BC-400, BC-408). The emission spectra of all six scintillators were obtained by using an optical spectrum analyzer and they were compared with theoretical behavior. For scintillation in the blue region, the signal intensity of a singly clad scintillating fiber (BCF-12) was 120% of that of the plastic scintillator (BC-400). For the multiclad fiber (SCSF-78), the signal reached 144% of that of the plastic scintillator. The intensity of the green scintillating fibers was lower than that of the plastic scintillator: 47% for the singly clad fiber (BCF-60) and 77% for the multiclad fiber (SCSF-3HF). The collected light was studied as a function of the scintillator length and radius for a cylindrical probe. We found that symmetric detectors with nearly the same spatial resolution in each direction (2 mm in diameter by 3 mm in length) could be made with a signal equivalent to those of the more commonly used asymmetric scintillators. With augmentation of the signal-to-noise ratio in consideration, this paper presents a series of comparisons that should provide insight into selection of a scintillator type and volume for development of a medical dosimeter

  20. Optimization of the scintillation parameters of the lead tungstate crystals for their application in high precision electromagnetic calorimetry

    International Nuclear Information System (INIS)

    Drobychev, G.

    2000-01-01

    In the frame of this dissertation work scintillation properties of the lead tungstate crystals PWO) and possibilities of their use were studied foreseeing their application for electromagnetic calorimetry in extreme radiation environment conditions of new colliders. The results of this work can be summarized in the following way. 1. A model of the scintillations origin in the lead tungstate crystals which includes processes influencing on the crystals radiation hardness and presence of slow components in scintillations was developed. 2. An analysis of the influences of the PWO scintillation properties changes on the parameters of the electromagnetic calorimeter was done. 3. Methods of the light collection from the large scintillation elements of complex shape made of the birefringent scintillation crystal with high refraction index and low light yield in case of signal registration by a photodetector with sensitive surface small in compare with the output face of scintillator were Studied. 4. Physical principles of the methodology of the scintillation crystals certification during their mass production foreseeing their installation into a calorimeter electromagnetic were developed. Correlations between the results of measurements of the PWO crystals parameters by different methods were found. (author)

  1. SBIR Final Report. Liquid Core Optical Scintillating Fibers

    International Nuclear Information System (INIS)

    Beetz, C.P.; Steinbeck, J.; Buerstler, R.

    2000-01-01

    This Phase I SBIR project focused on developing flexible scintillating liquid core optical fibers, with potential uses in high-energy calorimetry, tracking, preradiators, active targets or other fast detectors. Progress on the six tasks of the project is summarized. The technical developments involve three technology components: (1) highly flexible capillaries or tubes of relatively low n (index of refraction) to serve as cladding and liquid core containment; (2) scintillator (and clear) fluids of relatively high n to serve as a core-- these fluids must have a high light transmission and, for some applications, radiation hardness; (3) optical end plugs, plug insertion, and plug-cladding tube sealing technology to contain the core fluids in the tubes, and to transmit the light

  2. Evaluating scintillator performance in time-resolved hard X-ray studies at synchrotron light sources.

    Science.gov (United States)

    Rutherford, Michael E; Chapman, David J; White, Thomas G; Drakopoulos, Michael; Rack, Alexander; Eakins, Daniel E

    2016-05-01

    The short pulse duration, small effective source size and high flux of synchrotron radiation is ideally suited for probing a wide range of transient deformation processes in materials under extreme conditions. In this paper, the challenges of high-resolution time-resolved indirect X-ray detection are reviewed in the context of dynamic synchrotron experiments. In particular, the discussion is targeted at two-dimensional integrating detector methods, such as those focused on dynamic radiography and diffraction experiments. The response of a scintillator to periodic synchrotron X-ray excitation is modelled and validated against experimental data collected at the Diamond Light Source (DLS) and European Synchrotron Radiation Facility (ESRF). An upper bound on the dynamic range accessible in a time-resolved experiment for a given bunch separation is calculated for a range of scintillators. New bunch structures are suggested for DLS and ESRF using the highest-performing commercially available crystal LYSO:Ce, allowing time-resolved experiments with an interframe time of 189 ns and a maximum dynamic range of 98 (6.6 bits).

  3. Plastic Scintillators for Pulse Shape Discrimination of Particle Types in Radiation Detection

    Science.gov (United States)

    Hajagos, Tibor Jacob

    Organic scintillators have a long history in the field of radiation detection, dating back to some of the earliest studies of organic photophysics and optoelectronic properties. In particular, plastics have come to dominate the commercial market for organic scintillators, due to their low cost and ease of use and manufacturing, and more notably in spite of their poorer performance in many metrics. While there has been decades of active research since their inception, little progress has been made to improve upon the now well established compositions of commercial plastics, a notable exception being the recent development of plastic scintillators capable of pulse shape discrimination (PSD) of n/gamma radiation, which is of particular interest among governments and industry for the detection of illicit nuclear material and weapons. In recent years, much attention has been paid towards the study of luminescent organic materials, in particular due to the invention and widespread adoption of organic light emitting diode (OLED) based electronic devices, and the knowledge and lessons that have been fundamental to such fields have recently begun to be adopted by the organic scintilator community. In this work, new approaches to the design of both plastic scintillator components, and of the materials as a whole, are described, with particular emphasis paid towards the design and synthesis of small molecule scintillating dyes that are specifically tailored towards the development of PSD-capable plastic scintilators. In the first of these approaches, the design and synthesis of a highly soluble and polymerizable derivative of 9,10-diphenylanthracene is described, and the properties of plastic scintilators fabricated from this dye when copolymerized with poly(vinyl toluene) were investigated. This particular approach was used to demonstrate a proof-of-concept of PSD in highly loaded plastics stabilized through copolymerization of the primary dye, a strategy conceived to

  4. Scintillation camera

    International Nuclear Information System (INIS)

    Zioni, J.; Klein, Y.; Inbar, D.

    1975-01-01

    The scintillation camera is to make pictures of the density distribution of radiation fields created by the injection or administration radioactive medicaments into the body of the patient. It contains a scintillation crystal, several photomultipliers and computer circuits to obtain an analytical function at the exits of the photomultiplier which is dependent on the position of the scintillations at the time in the crystal. The scintillation crystal is flat and spatially corresponds to the production site of radiation. The photomultipliers form a pattern whose basic form consists of at least three photomultipliers. They are assigned to at least two crossing parallel series groups where a vertical running reference axis in the crystal plane belongs to each series group. The computer circuits are each assigned to a reference axis. Each series of a series group assigned to one of the reference axes in the computer circuit has an adder to produce a scintillation dependent series signal. Furthermore, the projection of the scintillation on this reference axis is calculated. A series signal is used for this which originates from a series chosen from two neighbouring photomultiplier series of this group. The scintillation must have appeared between these chosen series. They are termed as basic series. The photomultiplier can be arranged hexagonally or rectangularly. (GG/LH) [de

  5. On-line gross alpha radiation monitoring of natural waters with extractive scintillating resins

    International Nuclear Information System (INIS)

    Hughes, Lara; De Vol, T.A.

    2003-01-01

    Extractive scintillating resins, which are used to simultaneously separate and quantify radioactivity in aqueous solutions, were developed for low-level alpha radiation monitoring of natural waters. Resins were investigated with bis(2-ethylhexyl)methane-diphosphonic acid (H 2 DEH[MDP], Dipex[reg]) extractant, which has a strong affinity for tri-, tetra- and hexavalent actinides in dilute acids. Extractive scintillating resins were manifested (1) as a mixed bed of scintillating resin and extraction chromatographic resin and (2) by diffusing the organic fluor 2-(1-naphtyl)-5-phenyloxazole into macroporous polystyrene chromatographic resin, then coating with H 2 DEH[MDP], or by coating H 2 DEH[MDP] on scintillating polyvinyltoluene beads. The scintillation light was detected with a modified Hidex Triathler to allow for continuous flow measurements. The average detection efficiencies were 51.7±2.6% and 65.8±10.1% for natural uranium and 241 Am, respectively, for the extractant coated scintillator. The resin was stable for solution flow of up to 1000 ml resulting in rapid real-time quantification of natural uranium in groundwater down to 30 μg/ml

  6. Effect of gamma irradiation on optical properties of Ce 3+ - doped phosphate and silicate scintillating glasses

    Czech Academy of Sciences Publication Activity Database

    Baccaro, S.; Cecilia, A.; Chen, B.; Mareš, Jiří A.; Mihóková, Eva; Nikl, Martin; Polato, P.; Zanella, G.; Zannoni, R.

    2002-01-01

    Roč. 63, - (2002), s. 231-234 ISSN 0969-806X R&D Projects: GA MŠk ME 519 Institutional research plan: CEZ:AV0Z1010914 Keywords : glass scintillator * radiation hardness * colour centres Subject RIV: BM - Solid Matter Physics ; Magnetism Impact factor: 0.738, year: 2002

  7. Performance of molded plastic scintillators

    International Nuclear Information System (INIS)

    Gen, N.S.; Leman, V.E.; Solomonov, V.M.

    1989-01-01

    The performance of molded plastic scintillators is studied. The plastic scintillators studied were formed by transfer molding and intrusion from a scintillation composition consisting of polystyrene and a standard system of luminescent additives: 2 mass % of paraterphenyl + 0.06 mass % 1,4-di-/2-[5-phenyloxazoyly]/benzene and a plasticizer. The combined effect of mechanical load and temperature was studied. The effect of radiation on molded plastic scintillators was studied using gamma radiation from a 60 Co source. The studies show that the main operating characteristics of molded plastic scintillators are on a par with those of polymerized plastic scintillators. At the same time, molded plastic scintillators are superior in thermal stability at temperatures below the glass transition temperature and with respect to their working temperature range

  8. High-symmetry organic scintillator systems

    Science.gov (United States)

    Feng, Patrick L.

    2017-07-18

    An ionizing radiation detector or scintillator system includes a scintillating material comprising an organic crystalline compound selected to generate photons in response to the passage of ionizing radiation. The organic compound has a crystalline symmetry of higher order than monoclinic, for example an orthorhombic, trigonal, tetragonal, hexagonal, or cubic symmetry. A photodetector is optically coupled to the scintillating material, and configured to generate electronic signals having pulse shapes based on the photons generated in the scintillating material. A discriminator is coupled to the photon detector, and configured to discriminate between neutrons and gamma rays in the ionizing radiation based on the pulse shapes of the output signals.

  9. High-symmetry organic scintillator systems

    Energy Technology Data Exchange (ETDEWEB)

    Feng, Patrick L.

    2018-03-13

    An ionizing radiation detector or scintillator system includes a scintillating material comprising an organic crystalline compound selected to generate photons in response to the passage of ionizing radiation. The organic compound has a crystalline symmetry of higher order than monoclinic, for example an orthorhombic, trigonal, tetragonal, hexagonal, or cubic symmetry. A photodetector is optically coupled to the scintillating material, and configured to generate electronic signals having pulse shapes based on the photons generated in the scintillating material. A discriminator is coupled to the photon detector, and configured to discriminate between neutrons and gamma rays in the ionizing radiation based on the pulse shapes of the output signals.

  10. High-symmetry organic scintillator systems

    Science.gov (United States)

    Feng, Patrick L.

    2018-02-06

    An ionizing radiation detector or scintillator system includes a scintillating material comprising an organic crystalline compound selected to generate photons in response to the passage of ionizing radiation. The organic compound has a crystalline symmetry of higher order than monoclinic, for example an orthorhombic, trigonal, tetragonal, hexagonal, or cubic symmetry. A photodetector is optically coupled to the scintillating material, and configured to generate electronic signals having pulse shapes based on the photons generated in the scintillating material. A discriminator is coupled to the photon detector, and configured to discriminate between neutrons and gamma rays in the ionizing radiation based on the pulse shapes of the output signals.

  11. Radiation hard memory cell and array thereof

    International Nuclear Information System (INIS)

    Gunckel, T.L. II; Rovell, A.; Nielsen, R.L.

    1978-01-01

    A memory cell configuration that is implemented to be relatively hard to the adverse effects of a nuclear event is discussed. The presently disclosed memory cell can be interconnected with other like memory cells to form a high speed radiation hard register file. Information is selectively written into and read out of a memory cell comprising the register file, which memory cell preserves previously stored data without alteration in the event of exposure to high levels of nuclear radiation

  12. High spatial resolution radiation detectors based on hydrogenated amorphous silicon and scintillator

    International Nuclear Information System (INIS)

    Jing, T.; Lawrence Berkeley Lab., CA

    1995-05-01

    Hydrogenated amorphous silicon (a-Si:H) as a large-area thin film semiconductor with ease of doping and low-cost fabrication capability has given a new impetus to the field of imaging sensors; its high radiation resistance also makes it a good material for radiation detectors. In addition, large-area microelectronics based on a-Si:H or polysilicon can be made with full integration of peripheral circuits, including readout switches and shift registers on the same substrate. Thin a-Si:H p-i-n photodiodes coupled to suitable scintillators are shown to be suitable for detecting charged particles, electrons, and X-rays. The response speed of CsI/a-Si:H diode combinations to individual particulate radiation is limited by the scintillation light decay since the charge collection time of the diode is very short (< 10ns). The reverse current of the detector is analyzed in term of contact injection, thermal generation, field enhanced emission (Poole-Frenkel effect), and edge leakage. A good collection efficiency for a diode is obtained by optimizing the p layer of the diode thickness and composition. The CsI(Tl) scintillator coupled to an a-Si:H photodiode detector shows a capability for detecting minimum ionizing particles with S/N ∼20. In such an arrangement a p-i-n diode is operated in a photovoltaic mode (reverse bias). In addition, a p-i-n diode can also work as a photoconductor under forward bias and produces a gain yield of 3--8 for shaping times of 1 micros. The mechanism of the formation of structured CsI scintillator layers is analyzed. Initial nucleation in the deposited layer is sensitive to the type of substrate medium, with imperfections generally catalyzing nucleation. Therefore, the microgeometry of a patterned substrate has a significant effect on the structure of the CsI growth

  13. High spatial resolution radiation detectors based on hydrogenated amorphous silicon and scintillator

    Energy Technology Data Exchange (ETDEWEB)

    Jing, Tao [Univ. of California, Berkeley, CA (United States). Dept. of Engineering-Nuclear Engineering

    1995-05-01

    Hydrogenated amorphous silicon (a-Si:H) as a large-area thin film semiconductor with ease of doping and low-cost fabrication capability has given a new impetus to the field of imaging sensors; its high radiation resistance also makes it a good material for radiation detectors. In addition, large-area microelectronics based on a-Si:H or polysilicon can be made with full integration of peripheral circuits, including readout switches and shift registers on the same substrate. Thin a-Si:H p-i-n photodiodes coupled to suitable scintillators are shown to be suitable for detecting charged particles, electrons, and X-rays. The response speed of CsI/a-Si:H diode combinations to individual particulate radiation is limited by the scintillation light decay since the charge collection time of the diode is very short (< 10ns). The reverse current of the detector is analyzed in term of contact injection, thermal generation, field enhanced emission (Poole-Frenkel effect), and edge leakage. A good collection efficiency for a diode is obtained by optimizing the p layer of the diode thickness and composition. The CsI(Tl) scintillator coupled to an a-Si:H photodiode detector shows a capability for detecting minimum ionizing particles with S/N ~20. In such an arrangement a p-i-n diode is operated in a photovoltaic mode (reverse bias). In addition, a p-i-n diode can also work as a photoconductor under forward bias and produces a gain yield of 3--8 for shaping times of 1 {micro}s. The mechanism of the formation of structured CsI scintillator layers is analyzed. Initial nucleation in the deposited layer is sensitive to the type of substrate medium, with imperfections generally catalyzing nucleation. Therefore, the microgeometry of a patterned substrate has a significant effect on the structure of the CsI growth.

  14. Radionuclide identification algorithm for organic scintillator-based radiation portal monitor

    Energy Technology Data Exchange (ETDEWEB)

    Paff, Marc Gerrit, E-mail: mpaff@umich.edu; Di Fulvio, Angela; Clarke, Shaun D.; Pozzi, Sara A.

    2017-03-21

    We have developed an algorithm for on-the-fly radionuclide identification for radiation portal monitors using organic scintillation detectors. The algorithm was demonstrated on experimental data acquired with our pedestrian portal monitor on moving special nuclear material and industrial sources at a purpose-built radiation portal monitor testing facility. The experimental data also included common medical isotopes. The algorithm takes the power spectral density of the cumulative distribution function of the measured pulse height distributions and matches these to reference spectra using a spectral angle mapper. F-score analysis showed that the new algorithm exhibited significant performance improvements over previously implemented radionuclide identification algorithms for organic scintillators. Reliable on-the-fly radionuclide identification would help portal monitor operators more effectively screen out the hundreds of thousands of nuisance alarms they encounter annually due to recent nuclear-medicine patients and cargo containing naturally occurring radioactive material. Portal monitor operators could instead focus on the rare but potentially high impact incidents of nuclear and radiological material smuggling detection for which portal monitors are intended.

  15. Radionuclide identification algorithm for organic scintillator-based radiation portal monitor

    Science.gov (United States)

    Paff, Marc Gerrit; Di Fulvio, Angela; Clarke, Shaun D.; Pozzi, Sara A.

    2017-03-01

    We have developed an algorithm for on-the-fly radionuclide identification for radiation portal monitors using organic scintillation detectors. The algorithm was demonstrated on experimental data acquired with our pedestrian portal monitor on moving special nuclear material and industrial sources at a purpose-built radiation portal monitor testing facility. The experimental data also included common medical isotopes. The algorithm takes the power spectral density of the cumulative distribution function of the measured pulse height distributions and matches these to reference spectra using a spectral angle mapper. F-score analysis showed that the new algorithm exhibited significant performance improvements over previously implemented radionuclide identification algorithms for organic scintillators. Reliable on-the-fly radionuclide identification would help portal monitor operators more effectively screen out the hundreds of thousands of nuisance alarms they encounter annually due to recent nuclear-medicine patients and cargo containing naturally occurring radioactive material. Portal monitor operators could instead focus on the rare but potentially high impact incidents of nuclear and radiological material smuggling detection for which portal monitors are intended.

  16. BaYb2F8, a new radiation hard Cherenkov radiator for electromagnetic calorimeters

    International Nuclear Information System (INIS)

    Aseev, A.A.; Devitsin, E.G.; Komar, A.A.; Kozlov, V.A.; Hovsepyan, Yu.I.; Potashov, S.Yu.; Sokolovsky, K.A.; Uvarova, T.V.; Vasilchenko, V.G.

    1992-01-01

    Radiation hardness and optical properties of a new Cherenkov radiator, heavy fluoride BaYb 2 F 8 doped with various elements, have been studied. The above mentioned crystal has the density of 7 g/cm 3 , the radiation length is 1.28 cm and the Moliere radius 2.44 cm. High radiation hardness has been demonstrated for BaYb 2 F 8 doped with Tm, Pr, Tb. (orig.)

  17. Energy dependent response of plastic scintillation detectors to photon radiation of low to medium energy.

    Science.gov (United States)

    Ebenau, Melanie; Radeck, Désirée; Bambynek, Markus; Sommer, Holger; Flühs, Dirk; Spaan, Bernhard; Eichmann, Marion

    2016-08-01

    Plastic scintillation detectors are promising candidates for the dosimetry of low- to medium-energy photons but quantitative knowledge of their energy response is a prerequisite for their correct use. The purpose of this study was to characterize the energy dependent response of small scintillation detectors (active volume <1 mm(3)) made from the commonly used plastic scintillator BC400. Different detectors made from BC400 were calibrated at a number of radiation qualities ranging from 10 to 280 kV and at a (60)Co beam. All calibrations were performed at the Physikalisch-Technische Bundesanstalt, the National Metrology Institute of Germany. The energy response in terms of air kerma, dose to water, and dose to the scintillator was determined. Conversion factors from air kerma to dose to water and to dose to the scintillator were derived from Monte Carlo simulations. In order to quantitatively describe the energy dependence, a semiempirical model known as unimolecular quenching or Birks' formula was fitted to the data and from this the response to secondary electrons generated within the scintillator material BC400 was derived. The detector energy response in terms of air kerma differs for different scintillator sizes and different detector casings. It is therefore necessary to take attenuation within the scintillator and in the casing into account when deriving the response in terms of dose to water from a calibration in terms of air kerma. The measured energy response in terms of dose to water for BC400 cannot be reproduced by the ratio of mean mass energy-absorption coefficients for polyvinyl toluene to water but shows evidence of quenching. The quenching parameter kB in Birks' formula was determined to be kB = (12.3 ± 0.9) mg MeV(-1) cm(-2). The energy response was quantified relative to the response to (60)Co which is the common radiation quality for the calibration of therapy dosemeters. The observed energy dependence could be well explained with the

  18. Energy dependent response of plastic scintillation detectors to photon radiation of low to medium energy

    Energy Technology Data Exchange (ETDEWEB)

    Ebenau, Melanie, E-mail: melanie.ebenau@tu-dortmunde.de; Sommer, Holger; Spaan, Bernhard; Eichmann, Marion [Fakultät Physik, Technische Universität Dortmund, Otto-Hahn Str. 4a, 44221 Dortmund (Germany); Radeck, Désirée; Bambynek, Markus [Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig (Germany); Flühs, Dirk [Universitätsklinikum Essen, Hufelandstr. 55, 45147 Essen (Germany)

    2016-08-15

    Purpose: Plastic scintillation detectors are promising candidates for the dosimetry of low- to medium-energy photons but quantitative knowledge of their energy response is a prerequisite for their correct use. The purpose of this study was to characterize the energy dependent response of small scintillation detectors (active volume <1 mm{sup 3}) made from the commonly used plastic scintillator BC400. Methods: Different detectors made from BC400 were calibrated at a number of radiation qualities ranging from 10 to 280 kV and at a {sup 60}Co beam. All calibrations were performed at the Physikalisch-Technische Bundesanstalt, the National Metrology Institute of Germany. The energy response in terms of air kerma, dose to water, and dose to the scintillator was determined. Conversion factors from air kerma to dose to water and to dose to the scintillator were derived from Monte Carlo simulations. In order to quantitatively describe the energy dependence, a semiempirical model known as unimolecular quenching or Birks’ formula was fitted to the data and from this the response to secondary electrons generated within the scintillator material BC400 was derived. Results: The detector energy response in terms of air kerma differs for different scintillator sizes and different detector casings. It is therefore necessary to take attenuation within the scintillator and in the casing into account when deriving the response in terms of dose to water from a calibration in terms of air kerma. The measured energy response in terms of dose to water for BC400 cannot be reproduced by the ratio of mean mass energy-absorption coefficients for polyvinyl toluene to water but shows evidence of quenching. The quenching parameter kB in Birks’ formula was determined to be kB = (12.3 ± 0.9) mg MeV{sup −1} cm{sup −2}. Conclusions: The energy response was quantified relative to the response to {sup 60}Co which is the common radiation quality for the calibration of therapy dosemeters. The

  19. Further understanding of PbWO4 Scintillator characteristics and their optimisation. LUMEN activity in 1998

    CERN Document Server

    Baccaro, Stefania; Borgia, Bruno; Cecilia, Angelica; Dafinei, Ioan; Diemoz, Marcella; Fabeni, P; Festinesi, Armando; Longo, Egidio; Martini, M; Meinardi, F; Mihoková, E; Montecchi, Marco; Nikl, M; Pazzi, G P; Rosa, J; Sulc, Miroslav

    2000-01-01

    The aim of LUMEN collaboration was the investigation on single crystals of PbWO4 ( PWO): the results performed up to now provide the evidence of the possibility to optimise the optical properties of an intrinsic scintillator such as PWO. The control of essential requirements in the crystal preparation ( raw material purity, growing methods and post-growth annealing) as well as the introduction of selected dopants at suitable concentrations ( particularly trivalent and pentavalent ions) were found to be very successful in lowering the concentration of point defects in the lattice which strongly affect scintillation properties and radiation hardness. The systematic investigation effort to better understand the scintillation characteristics and to improve the quality of PWO crystals is due to their use for the CMS electromagnetic calorimeter.

  20. Scintillating Fibre Calorimetry at the LHC

    CERN Multimedia

    2002-01-01

    Good electromagnetic and hadronic calorimetry will play a central role in an LHC detector. The lead/scintillating fibre calorimeter technique provides a fast signal response well matched to the LHC rate requirements. It can be made to give equal response for electrons and hadrons (compensation) with good electromagnetic and hadronic energy resolutions.\\\\ \\\\ The aim of this R&D proposal is to study in detail the aspects that are relevant for application of this type of calorimeter in an LHC environment, including its integration in a larger system of detectors, e.g.~projective geometry, radiation hardness, light detection, calibration and stability monitoring, electron/hadron separation.....

  1. Physics of scintillation detectors

    International Nuclear Information System (INIS)

    Novotny, R.

    1991-01-01

    The general concept of a radiation detector is based on three fundamental principles: sensitivity of the device to the radiation of interest which requires a large cross-section in the detector material, detector response function to the physical properties of the radiation. As an example, a scintillation detector for charged particles should allow to identify the charge of the particle, its kinetic energy and the time of impact combined with optimum resolutions. Optimum conversion of the detector response (like luminescence of a scintillator) into electronical signals for further processing. The following article will concentrate on the various aspects of the first two listed principles as far as they appear to be relevant for photon and charged particle detection using organic and inorganic scintillation detectors. (orig.)

  2. arXiv Strong reduction of the effective radiation length in an oriented PWO scintillator crystal

    CERN Document Server

    Bandiera, L.; Romagnoni, M.; Argiolas, N.; Bagli, E.; Ballerini, G.; Berra, A.; Brizzolani, C.; Camattari, R.; De Salvador, D.; Haurylavets, V.; Mascagna, V.; Mazzolari, A.; Prest, M.; Soldani, M.; Sytov, A.; Vallazza, E.

    We measured a considerable increase of the emitted radiation by 120 GeV/c electrons in an axially oriented lead tungstate scintillator crystal, if compared to the case in which the sample was not aligned with the beam direction. This enhancement resulted from the interaction of particles with the strong crystalline electromagnetic field. The data collected at the external lines of CERN SPS were critically compared to Monte Carlo simulations based on the Baier Katkov quasiclassical method, highlighting a reduction of the scintillator radiation length by a factor of five in case of beam alignment with the [001] crystal axes. The observed effect opens the way to the realization of compact electromagnetic calorimeters/detectors based on oriented scintillator crystals in which the amount of material can be strongly reduced with respect to the state of the art. These devices could have relevant applications in fixed-target experiments as well as in satellite-borne gamma-telescopes.

  3. A scintillation method for measurement of the bioeffectiveness of ionizing radiation by simulation of the cellular response

    International Nuclear Information System (INIS)

    McDougall, I.C.; Alkharam, A.S.; Watt, D.E.; Thomas, G.E.

    1998-01-01

    A new type of radiation detector giving an absolute assessment of the bio-effectiveness of any ionizing radiation field is proposed. The spectrometer consists of a plastic scintillator disc 2.5 cm in diameter and 20 μm thick, interfaced to a hybrid photomultiplier. A close analogy exists between the major features of the track action in the scintillator and the track action which determines the radiation response in the mammalian cell. As the spectrometer should be capable to measure the equilibrium of charged particle fluence simultaneously, it has the potential of serving as a complete bioeffect dosemeter in a fluence-based system. (M.D.)

  4. Effect of gamma radiation on micromechanical hardness of lead-free solder joint

    Energy Technology Data Exchange (ETDEWEB)

    Paulus, Wilfred [Universiti Kebangsaan Malaysia, Bangi, 43600 Kajang, Selangor (Malaysia); Malaysian Nuclear Agency, Bangi, 43000 Kajang, Selangor (Malaysia); Rahman, Irman Abdul; Jalar, Azman; Kamil, Insan; Bakar, Maria Abu [Universiti Kebangsaan Malaysia, Bangi, 43600 Kajang, Selangor (Malaysia); Yusoff, Wan Yusmawati Wan [Universiti Pertahanan Nasional Malaysia, Kem Sg. Besi, 57000 Kuala Lumpur (Malaysia)

    2015-09-25

    Lead-free solders are important material in nano and microelectronic surface mounting technology for various applications in bio medicine, environmental monitoring, spacecraft and satellite instrumentation. Nevertheless solder joint in radiation environment needs higher reliability and resistance to any damage caused by ionizing radiations. In this study a lead-free 99.0Sn0.3Ag0.7Cu wt.% (SAC) solder joint was developed and subjected to various doses of gamma radiation to investigate the effects of the ionizing radiation to micromechanical hardness of the solder. Averaged hardness of the SAC joint was obtained from nanoindentation test. The results show a relationship between hardness values of indentations and the increment of radiation dose. Highest mean hardness, 0.2290 ± 0.0270 GPa was calculated on solder joint which was exposed to 5 Gray dose of gamma radiation. This value indicates possible radiation hardening effect on irradiated solder. The hardness gradually decreased to 0.1933 ± 0.0210 GPa and 0.1631 ± 0.0173 GPa when exposed to doses 50 and 500 gray respectively. These values are also lower than the hardness of non irradiated sample which was calculated as 0.2084 ± 0.0.3633 GPa indicating possible radiation damage and needs further related atomic dislocation study.

  5. The quest for the ideal inorganic scintillator

    International Nuclear Information System (INIS)

    Derenzo, S.E.; Weber, M.J.; Bourret-Courchesne, E.; Klintenberg, M.K.

    2002-01-01

    The past half century has witnessed the discovery of many new inorganic scintillator materials and numerous advances in our understanding of the basic physical processes governing the transformation of ionizing radiation into scintillation light. Whereas scintillators are available with a good combination of physical properties, none provides the desired combination of stopping power, light output, and decay time. A review of the numerous scintillation mechanisms of known inorganic scintillators reveals why none of them is both bright and fast. The mechanisms of radiative recombination in wide-bandgap direct semiconductors, however, remain relatively unexploited for scintillators. We describe how suitably doped semiconductor scintillators could provide a combination of high light output, short decay time, and linearity of response that approach fundamental limits

  6. Development of radiation hard components for ITER blanket remote handling system

    Energy Technology Data Exchange (ETDEWEB)

    Saito, Makiko, E-mail: saito.makiko@jaea.go.jp; Anzai, Katsunori; Maruyama, Takahito; Noguchi, Yuto; Ueno, Kenichi; Takeda, Nobukazu; Kakudate, Satoshi

    2016-11-01

    Highlights: • Clarify the components that will degrade by gamma ray irradiation. • Perform the irradiation tests to BRHS components. • Optimize the materials to increase the radiation hardness. - Abstract: The ITER blanket remote handling system (BRHS) will be operated in a high radiation environment (250 Gy/h max.) and must stably handle the blanket modules, which weigh 4.5 t and are more than 1.5 m in length, with a high degree of position and posture accuracy. The reliability of the system can be improved by reviewing the failure events of the system caused by high radiation. A failure mode and effects analysis (FMEA) identified failure modes and determined that lubricants, O-rings, and electric insulation cables were the dominant components affecting radiation hardness. Accordingly, we tried to optimize the lubricants and cables of the AC servo motors by using polyphenyl ether (PPE)-based grease and polyether ether ketone (PEEK), respectively. Materials containing radiation protective agents were also selected for the cable sheaths and O-rings to improve radiation hardness. Gamma ray irradiation tests were performed on these components and as a result, a radiation hardness of 8 MGy was achieved for the AC servo motors. On the other hand, to develop the radiation hardness and BRHS compatibility furthermore, the improvement of materials of cable and O ring were performed.

  7. Integration of Radiation-Hard Magnetic Random Access Memory with CMOS ICs

    CERN Document Server

    Cerjan, C J

    2000-01-01

    The research undertaken in this LDRD-funded project addressed the joint development of magnetic material-based nonvolatile, radiation-hard memory cells with Sandia National Laboratory. Specifically, the goal of this project was to demonstrate the intrinsic radiation-hardness of Giant Magneto-Resistive (GMR) materials by depositing representative alloy combinations upon radiation-hardened silicon-based integrated circuits. All of the stated goals of the project were achieved successfully. The necessary films were successfully deposited upon typical integrated circuits; the materials retained their magnetic field response at the highest radiation doses; and a patterning approach was developed that did not degrade the as-fabricated properties of the underlying circuitry. These results establish the feasibility of building radiation-hard magnetic memory cells.

  8. Photon statistics in scintillation crystals

    Science.gov (United States)

    Bora, Vaibhav Joga Singh

    Scintillation based gamma-ray detectors are widely used in medical imaging, high-energy physics, astronomy and national security. Scintillation gamma-ray detectors are eld-tested, relatively inexpensive, and have good detection eciency. Semi-conductor detectors are gaining popularity because of their superior capability to resolve gamma-ray energies. However, they are relatively hard to manufacture and therefore, at this time, not available in as large formats and much more expensive than scintillation gamma-ray detectors. Scintillation gamma-ray detectors consist of: a scintillator, a material that emits optical (scintillation) photons when it interacts with ionization radiation, and an optical detector that detects the emitted scintillation photons and converts them into an electrical signal. Compared to semiconductor gamma-ray detectors, scintillation gamma-ray detectors have relatively poor capability to resolve gamma-ray energies. This is in large part attributed to the "statistical limit" on the number of scintillation photons. The origin of this statistical limit is the assumption that scintillation photons are either Poisson distributed or super-Poisson distributed. This statistical limit is often dened by the Fano factor. The Fano factor of an integer-valued random process is dened as the ratio of its variance to its mean. Therefore, a Poisson process has a Fano factor of one. The classical theory of light limits the Fano factor of the number of photons to a value greater than or equal to one (Poisson case). However, the quantum theory of light allows for Fano factors to be less than one. We used two methods to look at the correlations between two detectors looking at same scintillation pulse to estimate the Fano factor of the scintillation photons. The relationship between the Fano factor and the correlation between the integral of the two signals detected was analytically derived, and the Fano factor was estimated using the measurements for SrI2:Eu, YAP

  9. Polycrystalline scintillators for large area detectors in HEP experiments

    Science.gov (United States)

    Dosovitskiy, G.; Fedorov, A.; Karpyuk, P.; Kuznetsova, D.; Mikhlin, A.; Kozlov, D.; Dosovitskiy, A.; Korjik, M.

    2017-06-01

    After significant increase of the accelerator luminosity throughout the High Luminosity phase of LHC, charged hadrons and neutrons with fluences higher than 1014 p/cm2 per year in the largest pseudo-rapidity regions of the detectors will cause increased radiation damage of materials. Increasing activation of the experimental equipment will make periodical maintenance and replacement of detector components difficult. Therefore, the selected materials for new detectors should be tolerant to radiation damage. Y3Al5O12:Ce (YAG:Ce) crystal was found to be one of the most radiation hard scintillation materials. However, production of YAG:Ce in a single crystalline form is costly, because crystal growth is performed at temperature near 1900°C with a very low rate of transformation of a raw material into a crystal. We propose translucent YAG:Ce ceramics as an alternative cheaper solution. Ceramic samples were sintered up to density ~98% of the theoretical value and were translucent. The samples have demonstrated light yield of 2200 phot./MeV under 662 keV γ-quanta, which gives the expected response to minimum ionizing particle around 3000 phot. for 2 mm thick plate. Scintillation light yield, registered under surface layer excitation with α-particles, was 50-70% higher than for the reference single crystal YAG:Ce.

  10. Scintillator material. Szintillatormaterial

    Energy Technology Data Exchange (ETDEWEB)

    Siegmund, M; Bendig, J; Regenstein, W

    1987-11-25

    A scintillator material for detection and quantitative determination of ionizing radiation is discussed consisting of an acridone dissolved in a fluid or solid medium. Solvent mixtures with at least one protogenic component or polymers and copolymers are used. The scintillator material is distinguished by an excellent stability at high energy doses.

  11. High efficiency scintillation detectors

    International Nuclear Information System (INIS)

    Noakes, J.E.

    1976-01-01

    A scintillation counter consisting of a scintillation detector, usually a crystal scintillator optically coupled to a photomultiplier tube which converts photons to electrical pulses is described. The photomultiplier pulses are measured to provide information on impinging radiation. In inorganic crystal scintillation detectors to achieve maximum density, optical transparency and uniform activation, it has been necessary heretofore to prepare the scintillator as a single crystal. Crystal pieces fail to give a single composite response. Means are provided herein for obtaining such a response with crystal pieces, such means comprising the combination of crystal pieces and liquid or solid organic scintillator matrices having a cyclic molecular structure favorable to fluorescence. 8 claims, 6 drawing figures

  12. WORKSHOP: Scintillating fibre detectors

    International Nuclear Information System (INIS)

    Anon.

    1989-01-01

    Scintillating fibre detector development and technology for the proposed US Superconducting Supercollider, SSC, was the subject of a recent workshop at Fermilab, with participation from the high energy physics community and from industry. Sessions covered the current status of fibre technology and fibre detectors, new detector applications, fluorescent materials and scintillation compositions, radiation damage effects, amplification and imaging structures, and scintillation fibre fabrication techniques

  13. Scintillation Counters

    Science.gov (United States)

    Bell, Zane W.

    Scintillators find wide use in radiation detection as the detecting medium for gamma/X-rays, and charged and neutral particles. Since the first notice in 1895 by Roentgen of the production of light by X-rays on a barium platinocyanide screen, and Thomas Edison's work over the following 2 years resulting in the discovery of calcium tungstate as a superior fluoroscopy screen, much research and experimentation have been undertaken to discover and elucidate the properties of new scintillators. Scintillators with high density and high atomic number are prized for the detection of gamma rays above 1 MeV; lower atomic number, lower-density materials find use for detecting beta particles and heavy charged particles; hydrogenous scintillators find use in fast-neutron detection; and boron-, lithium-, and gadolinium-containing scintillators are used for slow-neutron detection. This chapter provides the practitioner with an overview of the general characteristics of scintillators, including the variation of probability of interaction with density and atomic number, the characteristics of the light pulse, a list and characteristics of commonly available scintillators and their approximate cost, and recommendations regarding the choice of material for a few specific applications. This chapter does not pretend to present an exhaustive list of scintillators and applications.

  14. Radiation-hard silicon photonics for high energy physics and beyond

    CERN Multimedia

    CERN. Geneva

    2016-01-01

    Silicon photonics (SiPh) is currently being investigated as a promising technology for future radiation hard optical links. The possibility of integrating SiPh devices with electronics and/or silicon particle sensors as well as an expected very high resistance against radiation damage make this technology particularly interesting for potential use close to the interaction points in future in high energy physics experiments and other radiation-sensitive applications. The presentation will summarize the outcomes of the research on radiation hard SiPh conducted within the ICE-DIP projected.

  15. Comptonization of low-frequency radiation in accretion disks Angular distribution and polarization of hard X-ray radiation

    International Nuclear Information System (INIS)

    Suniaev, R.A.; Titarchuk, L.G.

    1984-01-01

    Analytical consideration is given to the comptonization of photons and its effects on the radiation emitted from accretion disks of compact X-ray sources, such as black holes and neutron stars. Attention is given to the photon distribution during escape from the disk, the angular distribution of hard radiation from the disk, the polarization of hard radiation and the electron temperature distribution over the optical depth. It is shown that the hard radiation spectrum is independent of the low-frequency photon source distribution. The angular distribution and polarization of the outgoing X-rays are a function of the optical depth. A Thomson approximation is used to estimate the angular distribution of the hard radiation and the polarization over the disk. The polarization results are compared with OSO-8 satellite data for Cyg X-1 and show good agreement at several energy levels. 17 references

  16. Functional possibilities of organosilicon coatings on the surface of CsI-based scintillators

    CERN Document Server

    Andryustchenko, L A; Goriletsky, V I; Zaslavsky, B G; Zosim, D I; Charkina, T A; Trefilova, L N; Renker, D; Ritt, S; Mzhavia, D A

    2002-01-01

    It has been shown that a thin film (15+-5 mu m) based on organosilicon coating applied to all surface of CsI and CsI(Tl) scintillators excluding the output window, can combine the following functions: (1) covering from atmospheric effects; (2) scintillation light convertor of luminescence towards the region of higher spectral sensitivity of the photoreceiver and (3) ancillary surface for performance of operations on changing the light collection coefficient without the risk to exceed limited size tolerations. Wavelength-shifting coating effect on radiation hardness of pure CsI is discussed. After irradiation a new absorption bands appear in the range 250-300 nm mainly. So, contrary to the 310 nm emission, the energy losses for converted light remain the same.

  17. Gamma spectrometry and plastic-scintillator inherent background

    International Nuclear Information System (INIS)

    Pomerantsev, V.V.; Gagauz, I.B.; Mitsai, L.I.; Pilipenko, V.S.; Solomonov, V.M.; Chernikov, V.V.; Tsirlin, Y.A.

    1988-01-01

    The authors measured the energy resolution for a linear dependence of light yield on gamma radiation energy of gamma spectrometers based on plastic scintillation detectors for several plastic scintillators. If there were several gamma lines from the source the line with the highest energy was used to eliminate distortion due to overlap from the Compton background from gamma radiation of higher energy. Attenuation lengths were calculated. The tests were based on three modes of interaction between the gamma radiation and the scintillator: Compton scattering, the photoelectric effect, and pair formation. The contribution from light collection was also considered. The scintillators tested included polystyrene, polymethyl methacrylate, cesium iodide, and sodium iodide. Gamma sources included cesium 137, sodium 22, potassium 40, yttrium 88, thorium 232, and plutonium-beryllium

  18. Scintillation detectors based on poly-2,4-dimethylstyrene: Structure peculiarities and radiation damage

    International Nuclear Information System (INIS)

    Gunder, O.A.; Voronkina, N.I.; Kopina, I.V.

    1995-01-01

    Scintillation detectors based on poly-2,4-dimethyl styrene (P-2,4-DMS) are studied. Investigated is the influence of two methyl groups present in the benzene ring on the energy, spectral and structural characteristics of the polymer. The said factors are assumed to result in the detectors high light output and radiation resistance. It is shown that under radiolysis (77 K) the radiation yield of the paramagnetic centers of P-2,4-DMS exceeds that of polystyrene (PST) by ∼ 1.5. Unlike PST film, the luminescence spectra of P-2,4-DMS are characterized by the presence of both excimer (320-340 nm) and monomer (292 nm) bands. Revealed are the distinction in the nature of the optical characteristics of macroradicals and the efficiency of energy transfer in gamma-irradiated PST and P-2,4-DMS scintillators. The relation between the super-molecular structure of the polymers and the interaction of their macroradicals with molecular O 2 is stated

  19. SPORT: A new sub-nanosecond time-resolved instrument to study swift heavy ion-beam induced luminescence - Application to luminescence degradation of a fast plastic scintillator

    Science.gov (United States)

    Gardés, E.; Balanzat, E.; Ban-d'Etat, B.; Cassimi, A.; Durantel, F.; Grygiel, C.; Madi, T.; Monnet, I.; Ramillon, J.-M.; Ropars, F.; Lebius, H.

    2013-02-01

    We developed a new sub-nanosecond time-resolved instrument to study the dynamics of UV-visible luminescence under high stopping power heavy ion irradiation. We applied our instrument, called SPORT, on a fast plastic scintillator (BC-400) irradiated with 27-MeV Ar ions having high mean electronic stopping power of 2.6 MeV/μm. As a consequence of increasing permanent radiation damages with increasing ion fluence, our investigations reveal a degradation of scintillation intensity together with, thanks to the time-resolved measurement, a decrease in the decay constant of the scintillator. This combination indicates that luminescence degradation processes by both dynamic and static quenching, the latter mechanism being predominant. Under such high density excitation, the scintillation deterioration of BC-400 is significantly enhanced compared to that observed in previous investigations, mainly performed using light ions. The observed non-linear behaviour implies that the dose at which luminescence starts deteriorating is not independent on particles' stopping power, thus illustrating that the radiation hardness of plastic scintillators can be strongly weakened under high excitation density in heavy ion environments.

  20. Radiation hardness of diamond and silicon sensors compared

    CERN Document Server

    de Boer, Wim; Furgeri, Alexander; Mueller, Steffen; Sander, Christian; Berdermann, Eleni; Pomorski, Michal; Huhtinen, Mika

    2007-01-01

    The radiation hardness of silicon charged particle sensors is compared with single crystal and polycrystalline diamond sensors, both experimentally and theoretically. It is shown that for Si- and C-sensors, the NIEL hypothesis, which states that the signal loss is proportional to the Non-Ionizing Energy Loss, is a good approximation to the present data. At incident proton and neutron energies well above 0.1 GeV the radiation damage is dominated by the inelastic cross section, while at non-relativistic energies the elastic cross section prevails. The smaller inelastic nucleon-Carbon cross section and the light nuclear fragments imply that at high energies diamond is an order of magnitude more radiation hard than silicon, while at energies below 0.1 GeV the difference becomes significantly smaller.

  1. Scintillation Detectors for Charged Particles and Photons

    CERN Document Server

    Lecoq, P

    2011-01-01

    Scintillation Detectors for Charged Particles and Photons in 'Charged Particle Detectors - Particle Detectors and Detector Systems', part of 'Landolt-Börnstein - Group I Elementary Particles, Nuclei and Atoms: Numerical Data and Functional Relationships in Science and Technology, Volume 21B1: Detectors for Particles and Radiation. Part 1: Principles and Methods'. This document is part of Part 1 'Principles and Methods' of Subvolume B 'Detectors for Particles and Radiation' of Volume 21 'Elementary Particles' of Landolt-Börnstein - Group I 'Elementary Particles, Nuclei and Atoms'. It contains the Subsection '3.1.1 Scintillation Detectors for Charged Particles and Photons' of Section '3.1 Charged Particle Detectors' of Chapter '3 Particle Detectors and Detector Systems' with the content: 3.1.1 Scintillation Detectors for Charged Particles and Photons 3.1.1.1 Basic detector principles and scintillator requirements 3.1.1.1.1 Interaction of ionizing radiation with scintillator material 3.1.1.1.2 Important scint...

  2. Solid scintillator 'Ready Cap' for measurement with a liquid scintillation counter

    International Nuclear Information System (INIS)

    Ijiri, Kenichi; Endo, Masashi; Nogawa, Norio; Tsuda, Shoko; Nakamura, Aiko; Morikawa, Naotake; Osaki, Susumu.

    1990-01-01

    'Ready Cap', a small plastic container coated with solid scintillator has recently been introduced (Beckman Instruments, Inc.). Pulse height spectra and counting efficiencies obtained with a liquid scintillator and Ready Cap using a liquid scintillation counter were compared for 15 different radionuclides. For radionuclides emitting low-energy β-rays or characteristic X-rays, the spectra for Ready Cap shifted toward the higher energy side compared with the spectra for the liquid scintillator. This tendency was reversed for the nuclides emitting higher-energy β-radiations ( 36 Cl and 32 P). Generally, counting efficiencies both in Ready Cap and in liquid scintillator increased with increase in the energy of β- or X-rays. For some nuclides, Ready Cap gave higher counting efficiencies and for others it gave lower values than in the liquid scintillator. However, the differences were not large within each nuclide. The use of Ready Cap is recommended for measurements of radionuclides when liquid scintillation cocktails have no means of waste disposal under the present Japanese radioisotope regulation. (author)

  3. Shutdown and degradation: Space computers for nuclear application, verification of radiation hardness. Final report

    International Nuclear Information System (INIS)

    Eichhorn, E.; Gerber, V.; Schreyer, P.

    1995-01-01

    (1) Employment of those radiation hard electronics which are already known in military and space applications. (2) The experience in space-flight shall be used to investigate nuclear technology areas, for example, by using space electronics to prove the range of applications in nuclear radiating environments. (3) Reproduction of a computer developed for telecommunication satellites; proof of radiation hardness by radiation tests. (4) At 328 Krad (Si) first failure of radiation tolerant devices with 100 Krad (Si) hardness guaranteed. (5) Using radiation hard devices of the same type you can expect applications at doses of greater than 1 Mrad (Si). Electronic systems applicable for radiation categories D, C and lower part of B for manipulators, vehicles, underwater robotics. (orig.) [de

  4. Plastic scintillator

    International Nuclear Information System (INIS)

    Andreeshchev, E.A.; Kilin, S.F.; Kavyrzina, K.A.

    1978-01-01

    A plastic scintillator for ionizing radiation detectors with high time resolution is suggested. To decrease the scintillation pulse width and to maintain a high light yield, the 4 1 , 4 5 -dibromo-2 1 , 2 5 , 5 1 , 5 5 -tetramethyl-n-quinquiphenyl (Br 2 Me 4 Ph) in combination with n-terphenyl (Ph 3 ) or 2, 5-diphenyloxadiazol-1, 3, 4 (PPD) is used as a luminescent addition. Taking into consideration the results of a special study, it is shown, that the following ratio of ingradients is the optimum one: 3-4 mass% Ph 3 or 4-7 mas% PPD + 2-5 mass% Br 2 Me 4 Ph + + polymeric base. The suggested scintillator on the basis of polystyrene has the light yield of 0.23-0.26 arbitrary units and the scintillation pulse duration at half-height is 0.74-0.84 ns

  5. Waveshifting fiber readout of lanthanum halide scintillators

    International Nuclear Information System (INIS)

    Case, G.L.; Cherry, M.L.; Stacy, J.G.

    2006-01-01

    Newly developed high-light-yield inorganic scintillators coupled to waveshifting optical fibers provide the capability of efficient X-ray detection and millimeter scale position resolution suitable for high-energy cosmic ray instruments, hard X-ray/gamma ray astronomy telescopes and applications to national security. The CASTER design for NASA's proposed Black Hole Finder Probe mission, in particular, calls for a 6-8 m 2 hard X-ray coded aperture imaging telescope operating in the 20-600 keV energy band, putting significant constraints on cost and readout complexity. The development of new inorganic scintillator materials (e.g., cerium-doped LaBr 3 and LaCl 3 ) provides improved energy resolution and timing performance that is well suited to the requirements for national security and astrophysics applications. LaBr 3 or LaCl 3 detector arrays coupled with waveshifting fiber optic readout represent a significant advance in the performance capabilities of scintillator-based gamma cameras and provide the potential for a feasible approach to affordable, large area, extremely sensitive detectors. We describe some of the applications and present laboratory test results demonstrating the expected scintillator performance

  6. Sci-Sat AM: Radiation Dosimetry and Practical Therapy Solutions - 10: Towards LET detection: A study on the effects of scintillator doping

    Energy Technology Data Exchange (ETDEWEB)

    Nusrat, Humza; Pang, Geordi; Ahmad, Syed; Keller, Brian; Sarfehnia, Arman [Ryerson University, Sunnybrook Health Sciences Centre (Canada)

    2016-08-15

    Purpose: In radiotherapy, the amount of radiation delivered is determined by optimizing the amount of absorbed dose to the tumor. Dose does not always correlate well with the actual biological effects of radiation. This work seeks to validate the LET-dependence of doped plastic scintillators for use in a radiation beam quality (LET) detector. Methods: The LET spectrum ([Φ]) can be resolved knowing the measured signals of uniquely LET-dependent detectors, [S], and the response of each LET-dependent detector to specific LETs ([R]), through the relation [Φ]=[S][R]{sup −1}. Plastic scintillator response is intrinsically LET dependent and can be varied via doping. Initial prototype consists of plastic scintillator and glass taper coupled to an optical fiber; components are housed in black acrylic, reducing effect of ambient light. In order to determine [R], the light response matrix, GEANT4.10.1 Monte Carlo (MC) was used. To validate MC, measurements were done using high energy electrons (9,12,15MeV) and orthovoltage x-rays (100,250kV); scintillator signal was normalized to dose measured simultaneously. Results: Stopping power was varied by changing particle type/energy; measurements indicated that as stopping power increased from 1.9 to 6.6MeV/cm, detector response increased by 263% (+/−29.2%) for 5%Pb-doped scintillator (155% in MC); 52% (+/−7.8%) increase observed when undoped scintillator was used (49% in MC). 5%Pb-doped discrepancy (100kV x-rays) is being investigated. Conclusions: This work validates that doping effects LET/energy response of scintillators; an effect that can be utilized for construction of an LET detector.

  7. Radioactivity measurements by liquid scintillation spectroscopy

    International Nuclear Information System (INIS)

    Cassette, Ph.

    2004-01-01

    The activity measurement techniques by liquid scintillation spectroscopy consist to mix the radioactive solution to measure with a scintillating liquid and to transform the ionizing radiations, resulting from decays, into light, detectable and quantifiable. The main advantages of these techniques are the easiness of preparation of the radioactive sources, the geometric efficiency of detection of 4π and the possibility of detection of low-level energy radiations. There are one of the only methods giving the possibility to measure the activity of pure β radionuclides; indeed, the nuclear disintegration is not accompanied of gamma radiations detectable by other techniques. There are one of the only methods too of measurement of radionuclides which disintegrate by electron capture and especially those leading to the emission of low-level energy ionizing radiations. Liquid scintillation spectroscopy can be used as an absolute method of activity measurement that is to say without the use of a calibration standard. The modern liquid scintillation counting devices can be very sensitive; the measurement of micro-activities being possible. Some of the applications of these activity measurement techniques are the carbon 14 dating and the geological tracing. Their main disadvantage is the global energetic yield which is low and variable in terms of the composition of the scintillation source necessitating to calculate the detection yield for each condition of measurement. (O.M.)

  8. Radiation hard diamond sensors for future tracking applications

    International Nuclear Information System (INIS)

    Adam, W.; Boer, W. de; Borchi, E.

    2006-01-01

    Progress in experimental particle physics in the coming decade depends crucially upon the ability to carry out experiments in high-radiation areas. In order to perform these complex and expensive experiments, new radiation hard technologies must be developed. This paper discusses the use of diamond detectors in future tracking applications and their survivability in the highest radiation environments. We present results of devices constructed with the newest polycrystalline and single crystal Chemical Vapor Deposition diamond and their tolerance to radiation

  9. Optical fiber composition and radiation hardness

    International Nuclear Information System (INIS)

    Wall, J.A.; Loretz, T.J.

    1982-01-01

    Germanium phosphosilicate and germanium borosilicate fibers doped with cerium were fabricated and tested for their responses to steady-state Co-60 radiation at -55 C, +20 C and +125 C. A fiber with germanium, boron and phosphorous in the silicate core and doped with antimony in the core and clad was similarly tested. All of the fibers showed significant improvements in radiation hardness at 20 C compared to undoped fibers of the same base composition. At -55 C, however, all except the cerium doped germanium phosphosilicate were very radiation sensitive and also showed increases in the rate of induced loss at +125 C. The cerium doped germanium phosphosilicate fiber showed virtually no change in radiation sensitivity at the temperature extremes and could prove useful in applications requiring relatively short lengths of fiber

  10. Semiconductor scintillator detector for gamma radiation; Detector cintilador semicondutor para radiacao gama

    Energy Technology Data Exchange (ETDEWEB)

    Laan, F.T.V. der; Borges, V.; Zabadal, J.R.S., E-mail: ftvdl@ufrgs.br, E-mail: borges@ufrgs.br, E-mail: jorge.zabadal@ufrgs.br [Universidade Federal do Rio Grande do Sul (GENUC/DEMEC/UFRGS), Porto Alegre, RS (Brazil). Grupo de Estudos Nucleares. Departamento de Engenharia Mecanica

    2015-07-01

    Nowadays the devices employed to evaluate individual radiation exposition are based on dosimetric films and thermoluminescent crystals, whose measurements must be processed in specific transductors. Hence, these devices carry out indirect measurements. Although a new generation of detectors based on semiconductors which are employed in EPD's (Electronic Personal Dosemeters) being yet available, it high producing costs and large dimensions prevents the application in personal dosimetry. Recent research works reports the development of new detection devices based on photovoltaic PIN diodes, which were successfully employed for detecting and monitoring exposition to X rays. In this work, we step forward by coupling a 2mm anthracene scintillator NE1, which converts the high energy radiation in visible light, generating a Strong signal which allows dispensing the use of photomultipliers. A low gain high performance amplifier and a digital acquisition device are employed to measure instantaneous and cumulative doses for energies ranging from X rays to Gamma radiation up to 2 MeV. One of the most important features of the PIN diode relies in the fact that it can be employed as a detector for ionization radiation, since it requires a small energy amount for releasing electrons. Since the photodiode does not amplify the corresponding photon current, it must be coupled to a low gain amplifier. Therefore, the new sensor works as a scintillator coupled with a photodiode PIN. Preliminary experiments are being performed with this sensor, showing good results for a wide range of energy spectrum. (author)

  11. Radiation Damage of the CERN CMS HCAL Scintillator/WLS fiber readout during Run1 and Run2 of the LHC

    CERN Document Server

    de Barbaro, Pawel Jan

    2017-01-01

    We present the results of a study of radiation damage of the CERN CMS HCAL Scintillator/WLS Fiber readout. Data were obtained using the Laser calibration system of the CMS hadron endcap detector during the operation of the LHC in 2010-2017. Scintillators used in the CMS hadron endcap calorimeter (HE) were irradiated at dose rates in the range of 0.1 rad/h to 0.1 krad/h. Results indicate that the radiation damage has a strong dose rate dependence. Using data collected in 2017, we have measured the response loss in a single HE section instrumented with Silicon photomultipliers (SiPMs). The results show a much smaller signal loss for the channels read out by SiPMs compared to signal loss for the channels read out by hybrid photodetectors (HPDs). The results imply that a large fraction of the response loss in the CMS HE detector observed in 2010-2017 comes from deterioration of the HPD photodetectors and not from radiation damage of scintillators.

  12. Design of Fluorescent Compounds for Scintillation Detection

    Energy Technology Data Exchange (ETDEWEB)

    Pla-Dalmau, Anna [Northern Illinois U.

    1990-01-01

    Plastic scintillation detectors for high energy physics applications require the development of new fluorescent compounds to meet the demands set by the future generation of particle accelerators such as the Superconducting Supercollider (SSe). Plastic scintillators are commonly based on a polymer matrix doped with two fluorescent compounds: the primary dopant and the wavelength shifter. Their main characteristics are fast response time and high quantum efficiency. The exposure to larger radiation doses and demands for larger light output questions their survivability in the future experiments. A new type of plastic scintillator - intrinsic scintillator - has been suggested. It uses a single dopant as primary and wavelength shifter, and should be less susceptible to radiation damage....

  13. Nanophosphor composite scintillator with a liquid matrix

    Science.gov (United States)

    McKigney, Edward Allen; Burrell, Anthony Keiran; Bennett, Bryan L.; Cooke, David Wayne; Ott, Kevin Curtis; Bacrania, Minesh Kantilal; Del Sesto, Rico Emilio; Gilbertson, Robert David; Muenchausen, Ross Edward; McCleskey, Thomas Mark

    2010-03-16

    An improved nanophosphor scintillator liquid comprises nanophosphor particles in a liquid matrix. The nanophosphor particles are optionally surface modified with an organic ligand. The surface modified nanophosphor particle is essentially surface charge neutral, thereby preventing agglomeration of the nanophosphor particles during dispersion in a liquid scintillator matrix. The improved nanophosphor scintillator liquid may be used in any conventional liquid scintillator application, including in a radiation detector.

  14. Use of internal scintillator radioactivity to calibrate DOI function of a PET detector with a dual-ended-scintillator readout

    International Nuclear Information System (INIS)

    Bircher, Chad; Shao Yiping

    2012-01-01

    Purpose: Positron emission tomography (PET) detectors that use a dual-ended-scintillator readout to measure depth-of-interaction (DOI) must have an accurate DOI function to provide the relationship between DOI and signal ratios to be used for detector calibration and recalibration. In a previous study, the authors used a novel and simple method to accurately and quickly measure DOI function by irradiating the detector with an external uniform flood source; however, as a practical concern, implementing external uniform flood sources in an assembled PET system is technically challenging and expensive. In the current study, therefore, the authors investigated whether the same method could be used to acquire DOI function from scintillator-generated (i.e., internal) radiation. The authors also developed a method for calibrating the energy scale necessary to select the events within the desired energy window. Methods: The authors measured the DOI function of a PET detector with lutetium yttrium orthosilicate (LYSO) scintillators. Radiation events originating from the scintillators' internal Lu-176 beta decay were used to measure DOI functions which were then compared with those measured from both an external uniform flood source and an electronically collimated external point source. The authors conducted these studies with several scintillators of differing geometries (1.5 x 1.5 and 2.0 x 2.0 mm 2 cross-section area and 20, 30, and 40 mm length) and various surface finishes (mirror-finishing, saw-cut rough, and other finishes in between), and in a prototype array. Results: All measured results using internal and external radiation sources showed excellent agreement in DOI function measurement. The mean difference among DOI values for all scintillators measured from internal and external radiation sources was less than 1.0 mm for different scintillator geometries and various surface finishes. Conclusions: The internal radioactivity of LYSO scintillators can be used to

  15. Experimental Study of the Lead Tungstate Scintillator Proton-Induced Damage and Recovery

    CERN Document Server

    Auffray, Etiennette; Singovski , A

    2011-01-01

    Lead tungstate (PbWO4, or PWO) scintillating crystals are used by two of the four experiments at the Large Hadron Collider (LHC): 75848 in CMS and 17920 in ALICE. For the CMS electromagnetic calorimeter, one of the most important crystal properties is its radiation hardness. With the increase of luminosity, the radiation level will increase drastically, particularly in the high pseudorapidity regions of the calorimeter. Beside the effects of color-centre formation caused by gamma-radiation, additional measurable effect originated by hadron irradiation could appear, which will further deteriorate the optical transmission of the crystals and therefore their efficiency. In this paper, we will present results of the proton-induced damage in PWO and a study of optical transmission recovery at different temperatures and under different light-induced "bleaching" conditions for proton-irradiated crystals.

  16. Liquid scintillation counting system with automatic gain correction

    International Nuclear Information System (INIS)

    Frank, R.B.

    1976-01-01

    An automatic liquid scintillation counting apparatus is described including a scintillating medium in the elevator ram of the sample changing apparatus. An appropriate source of radiation, which may be the external source for standardizing samples, produces reference scintillations in the scintillating medium which may be used for correction of the gain of the counting system

  17. Device for characterization of fissile materials comprising at least a neutron detector embedded inside a scintillator for gamma radiation detection

    International Nuclear Information System (INIS)

    Bernard, P.; Dherbey, J.R.; Bosser, R.; Berne, R.

    1989-01-01

    Fissile materials, for instance in radioactive wastes, are characterized by measurement of prompt and delayed neutrons and gamma radiation from induced fission by a neutron source. Gamma radiation is detected with a scintillation detector associated to a photomultiplier, the scintillation material is at the same time a moderator for thermalization of fast neutrons emitted by the neutron source and also of neutrons from spontaneous fission, (α, n) reactions and neutrons from induced fission in the fissile material. Preferentially the moderator is made of Altustipe (Plexiglas with anthracene as additive) [fr

  18. Phosphor scintillator structure

    International Nuclear Information System (INIS)

    Cusano, D.A.; Prener, J.S.

    1980-01-01

    A method of fabricating scintillators is described in which the phosphor is distributed within the structure in such a way as to enhance the escape of the visible wavelength radiation that would otherwise be dissipated within the scintillator body. Two embodiments of the present invention are disclosed: one in which the phosphor is distributed in a layered structure and another in which the phosphor is dispersed throughout a transparent matrix. (U.K.)

  19. Radiation-hard Silicon Photonics for Future High Energy Physics Experiments

    CERN Document Server

    AUTHOR|(CDS)2089774; Troska, Jan

    Collisions of proton beams in the Large Hadron Collider at CERN produce very high radiation levels in the innermost parts of the particle detectors and enormous amounts of measurement data. Thousands of radiation-hard optical links based on directly-modulated laser diodes are thus installed in the particle detectors to transmit the measurement data to the processing electronics. The radiation levels in the innermost regions of future particle detectors will be much higher than they are now. Alternative solutions to laser-based radiation-hard optical links have to be found since the performance of laser diodes decreases beyond the operation margin of the system when irradiated to sufficiently high radiation levels. Silicon Photonics (SiPh) is currently being investigated as a promising alternative technology. First tests have indeed shown that SiPh Mach-Zehnder modulators (MZMs) are relatively insensitive to a high neutron fluence. However, they showed a strong degradation when exposed to ionizing radiation. ...

  20. Scintillation of lead tungstate crystal studied with single-electron beam from KUFEL

    Energy Technology Data Exchange (ETDEWEB)

    Rizwan, Mohamad, E-mail: rizwan@nucl.kyushu-u.ac.jp; Uozumi, Yusuke; Matsuo, Kazuki [Department of Applied Quantum Physics and Nuclear Engineering, Kyushu University, Fukuoka (Japan); Ohgaki, Hideaki; Kii, Toshiteru; Zen, Heishun [Institute of Advanced Energy, Kyoto University, Gokasho, Uji, Kyoto (Japan); Tsamalaidze, Zviadi; Evtoukhovitch, Petr; Valentin, Samoilov [Joint Institute for Nuclear Research, JINR, Joliot-Curie Str.6, Dubna (Russian Federation)

    2015-04-29

    Lead tungstate (PWO) crystal has a very fast response, high atomic density and high radiation hardness. Therefore, they are suitable to be used for high-energy nuclear data measurements under high-background circumstances. Although a good electron-ion separation with a pulse shape analysis technique is essential, scintillation pulse shapes have not been observed with electron beams of a wide energy range. A single-electron beam technique has been developed at Kyoto University Free Electron Laser (KUFEL), and electron beams of 4-38 MeV are available. During the experiments, single electron beams bombarded a PWO crystal. By using oscilloscope we observed scintillation pulses of a PWO crystal coupled with a photomultiplier tube. Measured spectra were compared with the simulation code of EGS5 to analyze scattering effects. As the result, the pulse amplitudes show good linearity and the pulse shapes are almost constant in the observed energy range.

  1. Selective Beta and Gamma-ray Discrimination by CdWO{sub 4} and PlasticScintillator

    Energy Technology Data Exchange (ETDEWEB)

    Bae, Jun Woo; Kim, Hee Reyoung [Dept. of Nuclear Engineering, Ulsan National Institute of Science and Technology, Ulsan (Korea, Republic of)

    2017-04-15

    Radiation monitoring technique has been used for monitoring of decommissioning site of nuclear facility, radioactive waste disposal site, or in case of radioactivity accident. For rapid measurement of gamma-ray and beta-ray, many portable radiation detectors were developed but they are sensitive to specific radiation type. For example, portable detectors using NaI(Tl) or high purity germanium (HPGe) are suitable to detect gamma-ray. Otherwise, Geiger-müller (GM) tube or ionization chamber are suitable to detect all-types of radiation but it is hard to determine which particle is detected in the detector. In this reason, phoswich detectors for discrimination of beta-ray and gamma-ray were developed by using pulse shape discrimination. In this study, another approach to discriminate the beta-ray and gamma-ray is carried out. Two scintillators are used, cadmium tungstate (CdWO{sub 4}) and plastic scintillator. They have huge difference in their effective atomic number and mass density, thus they have huge difference in their gamma-ray sensitivity while the sensitivity of beta-ray is similar. The characterization of beta-ray and gamma-ray discrimination by using this characteristics is include. A technique of discrimination between beta-ray and gamma-ray was suggested. The method was verified by Monte Carlo simulation and experiment. This work showed feasibility on in field measurement of radiation with discrimination of beta-ray and gamma-ray.

  2. Building memristive and radiation hardness TiO{sub 2}-based junctions

    Energy Technology Data Exchange (ETDEWEB)

    Ghenzi, N., E-mail: n.ghenzi@gmail.com [Gerencia de Investigación y Aplicaciones, Comisión Nacional de Energía Atómica (Argentina); Rubi, D. [Gerencia de Investigación y Aplicaciones, Comisión Nacional de Energía Atómica (Argentina); ECyT, UNSAM, Martín de Irigoyen 3100, 1650 San Martín, Bs As (Argentina); Consejo Nacional de Investigaciones Científicas y Técnicas (CONICET) (Argentina); Mangano, E.; Gimenez, G. [Instituto Nacional de Tecnología Industrial (INTI) (Argentina); Lell, J. [Gerencia de Investigación y Aplicaciones, Comisión Nacional de Energía Atómica (Argentina); Zelcer, A. [Gerencia Química, Comisión Nacional de Energía Atómica (Argentina); ECyT, UNSAM, Martín de Irigoyen 3100, 1650 San Martín, Bs As (Argentina); Stoliar, P. [ECyT, UNSAM, Martín de Irigoyen 3100, 1650 San Martín, Bs As (Argentina); IMN, Université de Nantes, CNRS, 2 rue de la Houssinière, BP 32229, 44322 Nantes (France); and others

    2014-01-01

    We study micro-scale TiO{sub 2} junctions that are suitable to be used as resistive random-access memory nonvolatile devices with radiation hardness memristive properties. The fabrication and structural and electrical characterization of the junctions are presented. We obtained a retentivity of 10{sup 5} s, an endurance of 10{sup 4} cycles and reliable switching with short electrical pulses (time-width below 10 ns). Additionally, the devices were exposed to 25 MeV oxygen ions. Then, we performed electrical measurements comparing pristine and irradiated devices in order to check the feasibility of using these junctions as memory elements with memristive and radiation hardness properties. - Highlights: • We fabricated radiation hardness memristive metal insulator metal junctions. • We characterized the structural properties of the devices. • We showed the feasibility of the junctions as a non-volatile memory.

  3. Radiation hardness assurances categories for COTS technologies

    International Nuclear Information System (INIS)

    Hash, G.L.; Shaneyfelt, M.R.; Sexton, F.W.; Winokur, P.S.

    1997-01-01

    A comparison of the radiation tolerance of three commercial, and one radiation hardened SRAM is presented for four radiation environments. This work has shown the difficulty associated with strictly categorizing a device based solely on its radiation response, since its category depends on the specific radiation environment considered. For example, the 3.3-V Paradigm SRAM could be considered a radiation-tolerant device except for its SEU response. A more useful classification depends on the methods the manufacturer uses to ensure radiation hardness, i.e. whether specific design and process techniques have been used to harden the device. Finally, this work has shown that burned-in devices may fail functionally as much as 50% lower in total dose environments than non-burned-in devices. No burn-in effect was seen in dose-rate upset, latchup, or SEE environments. The user must ensure that total dose lot acceptance testing was performed on burned-in devices

  4. Radiation-damage studies, irradiations and high-dose dosimetry for LHC detectors

    CERN Document Server

    Coninckx, F; León-Florián, E; Leutz, H; Schönbacher, Helmut; Sonderegger, P; Tavlet, Marc; Sopko, B; Henschel, H; Schmidt, H U; Boden, A; Bräunig, D; Wulf, F; Cramariuc, R; Ilie, D; Fattibene, P; Onori, S; Miljanic, S; Paic, G; Razen, B; Razem, D; Rendic, D; CERN. Geneva. Detector Research and Development Committee

    1991-01-01

    The proposal is divided into a main project and special projects. The main project consists of a service similar to the one given in the past to accelerator construction projects at CERN (ISR,SPS,LEP) on high-dose dosimetry, material irradiations, irradiations tests, standardization of test procedures and data compilations. Large experience in this field and numerous radiation damage test data of insulating and structural materials are available. The special projects cover three topics which are of specific interest for LHC detector physicists and engineers at CERN and in other high energy physics institutes, namely: Radiation effects in scintillators; Selection of radiation hard optical fibres for data transmission; and Selection and testing of radiation hard electronic components.

  5. [Effects of ionizing radiation on scintillators and other particle detectors

    International Nuclear Information System (INIS)

    Proudfoot, J.

    1992-01-01

    It is my task to summarise the great variety of topics (covering a refreshing mix of physics, chemistry and technology) presented at this conference, which has focused on the effects of ionising radiation on scintillators and other particle detectors. One of the reasons and the central interest of many of the participants was the use of such detectors in experiments at two future large hadron colliders: the Superconducting Super Collider to be operating outside of Dallas in the United States by the turn of the decade and its European counterpart the Large Hadron Collider to be operating outside of Geneva in Switzerland on a similar time scale. These accelerators are the ''apple of the high energy physicist's eye.'' Their goal is to uncover the elusive Higgs particle and thereby set the cornerstone in our current knowledge of elementary particle interactions. This is the Quest, and from this lofty height the presentations rapidly moved on to the specific questions of experimental science: how such an experiment is carried out; why radiation damage is an issue; how radiation damage affects detectors; which factors affect radiation damage characteristics; which factors are not affected by radiation damage; and how better detectors may be constructed. These were the substance of this conference

  6. Search for new scintillators for high-energy resolution electromagnetic calorimeters

    International Nuclear Information System (INIS)

    Britvich, G.I.; Britvich, I.G.; Vasil'chenko, V.G.; Lishin, V.A.; Obraztsov, V.F.; Polyakov, V.A.; Solovjev, A.S.

    1999-01-01

    Some opportunities of creation of radiation-resistant heterogeneous electro-magnetic-calorimeters with an energy resolution of about σ/E≅4-5%/√E is given in this article. Investigation results of 2scintillation and radiation characteristics for thin molded plates and new heavy scintillators based on the polystyrene and containing metalloorganic additives are presented. The radiation resistance of thin molded scintillator plates of about 1.1 mm thick containing 2% pTP+0.05% POPOP has reached a level of about 15-20 kGy

  7. Development of new Polysiloxane Based Liquid Scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Dalla Palma, M.; Quaranta, A. [Department of Industrial Engineering, University of Trento,Via Sommarive, 9, 38123 Trento (Italy); INFN, Laboratori Nazionali di Legnaro,Viale dell' Universita, 2, 35020 Legnaro - Padova (Italy); Gramegna, F.; Marchi, T.; Cinausero, M. [INFN, Laboratori Nazionali di Legnaro,Viale dell' Universita, 2, 35020 Legnaro - Padova (Italy); Carturan, S.; Collazuol, G.; Checchia, C. [INFN, Laboratori Nazionali di Legnaro,Viale dell' Universita, 2, 35020 Legnaro - Padova (Italy); Department of Physics and Astronomy, University of Padova, Via Marzolo, 8, 35131 Padova (Italy); Degerlier, M. [Department of Physics, Nevsehir Haci Bektas Veli University, Science and Art Faculty, 50300 Nevsehir (Turkey)

    2015-07-01

    In the last decade, attention toward neutron detection has been growing in the scientific community, driven by new requirements in different fields of application ranging from homeland security to medical and material analysis, from research physics, to nuclear energy production. So far neutron detection, with particular attention to fast neutrons, has been mainly based on organic liquid scintillators, owing to their good efficiency and pulse shape discrimination (PSD) capability. Most of these liquids have however some main drawbacks given by toxicity, flammability, volatility and sensitivity to dissolved oxygen that limits the duration and the quality of their performances with worse handiness and increased costs. Phenyl-substituted polysiloxanes could address most of these issues, being characterized by low toxicity, low volatility and low flammability. Their optical properties can be tailored by changing the phenyl distribution and concentration thus allowing to increase the solubility of organic dyes, to modify the fluorescence spectra and to vary the refractive index of the medium. Furthermore, polysiloxanes have been recently exploited for the production of plastic scintillators with very good chemical and thermal stability and very good radiation hardness and the development of polysiloxane liquid scintillators could allow to combine these interesting properties with the supremacy of liquid scintillators as regarding PSD over plastics. For these reasons, the properties of several phenyl-substituted polysiloxane with different phenyl amounts and different viscosities have been investigated, with particular attention to the scintillation response and the pulse shape discrimination capability, and the results of the investigation are reported in this work. More in details, the scintillation light yield towards gamma rays ({sup 60}Co and {sup 137}Cs) of several polysiloxane liquids has been analyzed and compared with the light yield of a commercial non

  8. Impact of precursor purity on optical properties and radiation detection of CsI:Tl scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Saengkaew, Phannee; Cheewajaroen, Kulthawat; Yenchai, Chadet; Thong-aram, Decho [Chulalongkorn University, Department of Nuclear Engineering, Faculty of Engineering, Bangkok (Thailand); Sanorpim, Sakuntam [Chulalongkorn University, Department of Physics, Faculty of Science, Bangkok (Thailand); Jitpukdee, Manit [Kasetsart University, Department of Applied Radiation and Isotope, Faculty of Science, Bangkok (Thailand); Yordsri, Visittapong; Thanachayanont, Chanchana [Ministry of Science and Technology, National Metal and Materials Technology Center, National Science and Technology Development Agency, Pathumthani (Thailand); Nuntawong, Noppadon [Ministry of Science and Technology, National Electronic and Computer Technology Center, National Science and Technology Development Agency, Pathumthani (Thailand)

    2016-08-15

    Cesium iodide doped with thallium (CsI:Tl) crystals was grown to develop the gamma-ray detectors by using low-cost raw materials. Effect of impurities on optical properties and radiation detection performance was investigated. By a modified homemade Bridgman-Stockbarger technique, CsI:Tl samples were grown in two levels of CsI and TlI reactant materials, i.e., having as a very high purity of 99.999 % and a high purity of 99.9 %. XRD measurements indicate CsI:Tl crystals having a good quality with a dominant (110) plane. Having a cubic structure, a lattice constant of CsI crystals of 0.4574 nm and a crystallite size of 43.539 nm were obtained. From the lower-purity raw materials, calcite was found in an orange crystal with a lattice constant of 0.4560 nm and a crystallite size of 43.089 nm. By PL measurements, the optical properties of the CsI:Tl crystals were analyzed. ∝540-nm-wavelength PL peak was observed from the colorless high-purity crystal, and ∝600-nm-wavelength PL peak was observed from the orange crystal. The brighter PL emission was obtained from the orange crystals suggesting impurities. CsI:Tl surface morphology by SEM exhibited a smooth surface with some parallel crystal facets. For electrical properties of high-quality CsI:Tl crystals, the electrical resistances were 230 ± 16 MΩ in cross-sectional direction and 714 ± 136 MΩ in vertical direction with respect to more homogeneous crystal quality in cross-sectional direction than that in vertical direction. TEM measurement was applied to evaluate the microstructure of colorless CsI:Tl crystal with different patterns of a cubic structure. Both CsI:Tl crystals show good efficiencies and good resolutions. Maintaining the same electronic conditions and amplifications, the colorless CsI:Tl scintillators represented a higher detection efficiency at 122 keV of Co-57 of 78.4 % and the energy resolution of 23.3 % compared to the detection efficiency of 75.9 % and the energy resolution of 34.6 % of the

  9. Optimized mounting of a polyethylene naphthalate scintillation material in a radiation detector.

    Science.gov (United States)

    Nakamura, Hidehito; Yamada, Tatsuya; Shirakawa, Yoshiyuki; Kitamura, Hisashi; Shidara, Zenichiro; Yokozuka, Takayuki; Nguyen, Philip; Kanayama, Masaya; Takahashi, Sentaro

    2013-10-01

    Polyethylene naphthalate (PEN) has great potential as a scintillation material for radiation detection. Here the optimum mounting conditions to maximize the light collection efficiency from PEN in a radiation detector are discussed. To this end, we have determined light yields emitted from irradiated PEN for various optical couplings between the substrate and the photodetector, and for various substrate surface treatments. The results demonstrate that light extraction from PEN is more sensitive to the optical couplings due to its high refractive index. We also assessed the extent of radioactive impurities in PEN as background sources and found that the impurities are equivalent to the environmental background level. Copyright © 2013 Elsevier Ltd. All rights reserved.

  10. Comparative study using Monte Carlo methods of the radiation detection efficiency of LSO, LuAP, GSO and YAP scintillators for use in positron emission imaging (PET)

    International Nuclear Information System (INIS)

    Nikolopoulos, Dimitrios; Kandarakis, Ioannis; Tsantilas, Xenophon; Valais, Ioannis; Cavouras, Dionisios; Louizi, Anna

    2006-01-01

    The radiation detection efficiency of four scintillators employed, or designed to be employed, in positron emission imaging (PET) was evaluated as a function of the crystal thickness by applying Monte Carlo Methods. The scintillators studied were the LuSiO 5 (LSO), LuAlO 3 (LuAP), Gd 2 SiO 5 (GSO) and the YAlO 3 (YAP). Crystal thicknesses ranged from 0 to 50 mm. The study was performed via a previously generated photon transport Monte Carlo code. All photon track and energy histories were recorded and the energy transferred or absorbed in the scintillator medium was calculated together with the energy redistributed and retransported as secondary characteristic fluorescence radiation. Various parameters were calculated e.g. the fraction of the incident photon energy absorbed, transmitted or redistributed as fluorescence radiation, the scatter to primary ratio, the photon and energy distribution within each scintillator block etc. As being most significant, the fraction of the incident photon energy absorbed was found to increase with increasing crystal thickness tending to form a plateau above the 30 mm thickness. For LSO, LuAP, GSO and YAP scintillators, respectively, this fraction had the value of 44.8, 36.9 and 45.7% at the 10 mm thickness and 96.4, 93.2 and 96.9% at the 50 mm thickness. Within the plateau area approximately (57-59)% (59-63)% (52-63)% and (58-61)% of this fraction was due to scattered and reabsorbed radiation for the LSO, GSO, YAP and LuAP scintillators, respectively. In all cases, a negligible fraction (<0.1%) of the absorbed energy was found to escape the crystal as fluorescence radiation

  11. Development of ITER diagnostics: Neutronic analysis and radiation hardness

    Energy Technology Data Exchange (ETDEWEB)

    Vukolov, Konstantin, E-mail: vukolov_KY@nrcki.ru; Borisov, Andrey; Deryabina, Natalya; Orlovskiy, Ilya

    2015-10-15

    Highlights: • Problems of ITER diagnostics caused by neutron radiation from hot DT plasma considered. • Careful neutronic analysis is necessary for ITER diagnostics development. • Effective nuclear shielding for ITER diagnostics in the 11th equatorial port plug proposed. • Requirements for study of radiation hardness of diagnostic elements defined. • Results of optical glasses irradiation tests in a fission reactor given. - Abstract: The paper is dedicated to the problems of ITER diagnostics caused by effects of radiation from hot DT plasma. An effective nuclear shielding must be arranged in diagnostic port plugs to meet the nuclear safety requirements and to provide reliable operation of the diagnostics. This task can be solved with the help of neutronic analysis of the diagnostics environment within the port plugs at the design stage. Problems of neutronic calculations are demonstrated for the 11th equatorial port plug. The numerical simulation includes the calculations of neutron fluxes in the port-plug and in the interspace. Options for nuclear shielding, such as tungsten collimator, boron carbide and water moderators, stainless steel and lead screens are considered. Data on neutron fluxes along diagnostic labyrinths allow to define radiation hardness requirements for the diagnostic components and to specify their materials. Options for windows and lenses materials for optical diagnostics are described. The results of irradiation of flint and silica glasses in nuclear reactor have shown that silica KU-1 and KS-4V retain transparency in visible range after neutron fluence of 10{sup 17} cm{sup −2}. Flints required for achromatic objectives have much less radiation hardness about 5 × 10{sup 14} n/cm{sup 2}.

  12. Development of scintillators on the basis of AIIBVI compounds for radiation instruments used in medical and technical applications

    Directory of Open Access Journals (Sweden)

    Starzhinskiy N. G.

    2009-06-01

    Full Text Available Physico-technological problems of preparation and main properties of scintillation materials based on zinc selenide and other AIIBVI compounds are considered. Effects have been determined of dopant properties on formation processes of complex lattice defects playing the role of luminescence centers. It is shown that such property features as high light output and very low afterglow level, as well as a unique combination of scintillation and semiconductor properties allow application of these materials in different fields of radiation instrument technologies.

  13. Investigation of optically-induced DFB lasing in organic scintillators in order to achieve active ionizing radiation measurement

    International Nuclear Information System (INIS)

    Michel, Maugan

    2014-01-01

    Transducers used for nuclear measurements are divided into two groups. The first one is made of detectors based on the movement of charged carriers created during the interaction of ionizing radiation with materials (ionization chambers, semiconductors, etc.). The second comprises scintillating materials which emit light following such interaction. The present work aims at shifting the current paradigm in order to achieve an active measurement of the interactions between particles and sensor. The aim of this thesis is to investigate the response of nano-structured scintillators when excited by a primary laser beam while in the presence of ionizing radiations. It consists more precisely in optically inducing a laser emission in scintillators so as to benefit from the inherent sensitivity of this kind of wave to any kind of change in its environment. This study is at first focused on controlling the propagation of the electromagnetic waves in nano-structured media. This nano-structuration allows the channeling of the electromagnetic energy in a particular direction and at a precise wavelength and thus to amplify the transduction signal. The second part of this study is dedicated to the attempts at observing any change in the emission in an amplified regime while exposed to ionizing radiations (alpha, beta). The limitation of these sources requires the use of an electron accelerator. The conclusion of this part lays the foundation for future work and provides hypotheses regarding the effects that could be expected from the irradiation with high electron fluxes. (author) [fr

  14. Extreme Radiation Hardness and Space Qualification of AlGaN Optoelectronic Devices

    International Nuclear Information System (INIS)

    Sun, Ke-Xun; MacNeil, Lawrence; Balakrishnan, Kathik; Hultgren, Eric; Goebel, John; Bilenko, Yuri; Yang, Jinwei; Sun, Wenhong; Shatalov, Max; Hu, Xuhong; Gaska, Remis

    2010-01-01

    Unprecedented radiation hardness and environment robustness are required in the new generation of high energy density physics (HEDP) experiments and deep space exploration. National Ignition Facility (NIF) break-even shots will have a neutron yield of 10 15 or higher. The Europa Jupiter System Mission (EJSM) mission instruments will be irradiated with a total fluence of 10 12 protons/cm 2 during the space journey. In addition, large temperature variations and mechanical shocks are expected in these applications under extreme conditions. Hefty radiation and thermal shields are required for Si and GaAs based electronics and optoelectronics devices. However, for direct illumination and imaging applications, shielding is not a viable option. It is an urgent task to search for new semiconductor technologies and to develop radiation hard and environmentally robust optoelectronic devices. We will report on our latest systematic experimental studies on radiation hardness and space qualifications of AlGaN optoelectronic devices: Deep UV Light Emitting Diodes (DUV LEDs) and solarblind UV Photodiodes (PDs). For custom designed AlGaN DUV LEDs with a central emission wavelength of 255 nm, we have demonstrated its extreme radiation hardness up to 2 x 10 12 protons/cm 2 with 63.9 MeV proton beams. We have demonstrated an operation lifetime of over 26,000 hours in a nitrogen rich environment, and 23,000 hours of operation in vacuum without significant power drop and spectral shift. The DUV LEDs with multiple packaging styles have passed stringent space qualifications with 14 g random vibrations, and 21 cycles of 100K temperature cycles. The driving voltage, current, emission spectra and optical power (V-I-P) operation characteristics exhibited no significant changes after the space environmental tests. The DUV LEDs will be used for photoelectric charge management in space flights. For custom designed AlGaN UV photodiodes with a central response wavelength of 255 nm, we have

  15. Detection of gamma rays using scintillation optical fibers

    International Nuclear Information System (INIS)

    Park, J. W.; Hong, S. B.

    2002-01-01

    Scintillating optical fibers have several advantages over other conventional materials used for radiation detection. We have used glass and plastic scintillating fibers to detect gamma rays emitted from 60 Co and 137 Cs, and beta rays from 90 Sr. The sensors are constructed of single strand or multi-strand fibers of 1 mm diameter. The glass scintillating fiber used contains cerium-activated lithium-silicate as scintillating material and the plastic scintillating fiber used is Bicron model BCF-12. In this paper, we report the pulse-height spectra obtained by both sensor types, and analyze them in the aspect of their usability for radiation detectors. Our investigation suggests that the glass fiber can be used to develop gamma ray detectors which will function in high and low gamma ray flux environments. Use of the sensor for the beta ray detection was not satisfactory. The plastic fiber sensor did not work satisfactorily for the weak gamma sources, but did produce somewhat promising results. The scintillating plastic fiber offers some feasibility as beta ray sensor material

  16. Development of Radiation Hard Radiation Detectors, Differences between Czochralski Silicon and Float Zone Silicon

    CERN Document Server

    Tuominen, Eija

    2012-01-01

    The purpose of this work was to develop radiation hard silicon detectors. Radiation detectors made ofsilicon are cost effective and have excellent position resolution. Therefore, they are widely used fortrack finding and particle analysis in large high-energy physics experiments. Silicon detectors willalso be used in the CMS (Compact Muon Solenoid) experiment that is being built at the LHC (LargeHadron Collider) accelerator at CERN (European Organisation for Nuclear Research). This work wasdone in the CMS programme of Helsinki Institute of Physics (HIP).Exposure of the silicon material to particle radiation causes irreversible defects that deteriorate theperformance of the silicon detectors. In HIP CMS Programme, our approach was to improve theradiation hardness of the silicon material with increased oxygen concentration in silicon material. Westudied two different methods: diffusion oxygenation of Float Zone silicon and use of high resistivityCzochralski silicon.We processed, characterised, tested in a parti...

  17. Measuring the emulsion stability in Cherenkov radiation with insignificant modification of a liquid scintillation spectrometer

    International Nuclear Information System (INIS)

    Wiechen, A.; Lorenzen, P.Ch.; Reimerdes, E.H.

    1984-01-01

    A method is described by which the stability of emulsions can be measured by a modified liquid scintillation counter. The 226 Ra external standard source of a commercially available equipment, fixed in the measuring position, is used for the production of Cherenkov radiation in a sample of an emulsion. This Cherenkov radiation is absorbed by the sample due to its turbidity. The turbidity of emulsions follows a typical course with time designated as creaming-up-curve. These curves can be registered automatically in digital form. (author)

  18. Set of counts by scintillations for atmospheric samplings; Ensemble de comptages par scintillations pour prelevements atmospheriques

    Energy Technology Data Exchange (ETDEWEB)

    Appriou, D.; Doury, A.

    1962-07-01

    The author reports the development of a scintillation-based counting assembly with the following characteristics: a photo-multiplier with a wide photo-cathode, a thin plastic scintillator for the counting of beta + alpha (and possibility of mounting an alpha scintillator), a relatively small own motion with respect to activities to be counted, a weakly varying efficiency. The authors discuss the counting objective, present equipment tests (counter, proportional amplifier and pre-amplifier, input drawer). They describe the apparatus operation, discuss the selection of scintillators, report the study of the own movement (electron-based background noise, total background noise, background noise reduction), discuss counts (influence of the external source, sensitivity to alpha radiations, counting homogeneity, minimum detectable activity) and efficiencies.

  19. Chronic radiation effects on dental hard tissue (''radiation carries''). Classification and therapeutic strategies

    International Nuclear Information System (INIS)

    Groetz, K.A.; Brahm, R.; Al-Nawas, B.; Wagner, W.; Riesenbeck, D.; Willich, N.; Seegenschmiedt, M.H.

    2001-01-01

    Objectives: Since the first description of rapid destruction of dental hard tissues following head and neck radiotherapy 80 years ago, 'radiation caries' is an established clinical finding. The internationally accepted clinical evaluation score RTOG/EORTC however is lacking a classification of this frequent radiogenic alteration. Material and Methods: Medical records, data and images of radiation effects on the teeth of more than 1,500 patients, who underwent periradiotherapeutic care, were analyzed. Macroscopic alterations regarding the grade of late lesions of tooth crowns were used for a classification into 4 grades according to the RTOG/EORTC guidelines. Results: No early radiation effects were found by macroscopic inspection. In the first 90 days following radiotherapy 1/3 of the patients complained of reversible hypersensitivity, which may be related to a temporary hyperemia of the pulp. It was possible to classify radiation caries as a late radiation effect on a graded scale as known from RTOG/EORTC for other organ systems. This is a prerequisite for the integration of radiation caries into the international nomenclature of the RTOG/EORTC classification. Conclusions: The documentation of early radiation effects on dental hard tissues seems to be neglectable. On the other hand the documentation of late radiation effects has a high clinical impact. The identification of an initial lesion at the high-risk areas of the neck and incisal part of the tooth can lead to a successful therapy as a major prerequisite for orofacial rehabilitation. An internationally standardized documentation is a basis for the evaluation of the side effects of radiooncotic therapy as well as the effectiveness of protective and supportive procedures. (orig.) [de

  20. Scintillating plate calorimeter optical design

    International Nuclear Information System (INIS)

    McNeil, R.; Fazely, A.; Gunasingha, R.; Imlay, R.; Lim, J.

    1990-01-01

    A major technical challenge facing the builder of a general purpose detector for the SSC is to achieve an optimum design for the calorimeter. Because of its fast response and good energy resolution, scintillating plate sampling calorimeters should be considered as a possible technology option. The work of the Scintillating Plate Calorimeter Collaboration is focused on compensating plate calorimeters. Based on experimental and simulation studies, it is expected that a sampling calorimeter with alternating layers of high-Z absorber (Pb, W, DU, etc.) and plastic scintillator can be made compensating (e/h = 1.00) by suitable choice of the ratio of absorber/scintillator thickness. Two conceptual designs have been pursued by this subsystem collaboration. One is based on lead as the absorber, with read/out of the scintillator plates via wavelength shifter fibers. The other design is based on depleted uranium as the absorber with wavelength shifter (WLS) plate readout. Progress on designs for the optical readout of a compensating scintillator plate calorimeter are presented. These designs include readout of the scintillator plates via wavelength shifter plates or fiber readout. Results from radiation damage studies of the optical components are presented

  1. Improvements to well scintillation counters

    International Nuclear Information System (INIS)

    Farukhi, M.R.; Mataraza, G.A.; Wimer, O.D.

    1977-01-01

    This invention relates to the field of ionising radiation detection. It concerns in particular scintillation detectors of the type that is commonly used in conjunction with a photomultiplier tube and that is used for monitoring radiation, for instance in the clinical measurements of isotopes. This invention enables well scintillation counters to be made, characterised by a high efficiency in measuring the thindown rate of radio-pharmaceutical solutions and to resolve the distribution of energy emanating from the radioactive source. It particularly consists in improving the uniformity of the luminous efficiency, the quality of the resolution and the efficiency whilst improving the reception of light [fr

  2. Plastic scintillators with {beta}-diketone Eu complexes for high ionizing radiation detection

    Energy Technology Data Exchange (ETDEWEB)

    Adadurov, A.F., E-mail: adadurov@isma.kharkov.ua [Institute for Scintillating Materials, NAN of Ukraine, Lenin Avenue 60, 61001 Kharkov (Ukraine); Zhmurin, P.N.; Lebedev, V.N.; Kovalenko, V.N. [Institute for Scintillating Materials, NAN of Ukraine, Lenin Avenue 60, 61001 Kharkov (Ukraine)

    2011-10-15

    Luminescent and scintillation properties of polystyrene-based plastic scintillators with {beta}-diketone Eu complexes are investigated. A scintillator with dibenzoylmethane Eu complex containing two phenyl groups demonstrates the maximum scintillating efficiency. It is shown that plastic scintillators efficiency is dramatically decreased if {beta}-diketone derivatives contain no phenyl groups as substituents. This fact can be explained by exciplex mechanism of energy transfer from a matrix to Eu complex. - Highlights: > Fluorescent properties of polystyrene scintillators with {beta}-diketone complexes of Eu were studied. > Scintillating efficiency is increased with the number of phenyl groups in Eu complex. > This is related to exciplex mechanism of energy transfer from a polymer matrix to Eu complex.

  3. Transition radiation in metal-metal multilayer nanostructures as a medical source of hard x-ray radiation

    International Nuclear Information System (INIS)

    Pokrovsky, A. L.; Kaplan, A. E.; Shkolnikov, P. L.

    2006-01-01

    We show that a periodic metal-metal multilayer nanostructure can serve as an efficient source of hard x-ray transition radiation. Our research effort is aimed at developing an x-ray source for medical applications, which is based on using low-energy relativistic electrons. The approach toward choosing radiator-spacer couples for the generation of hard x-ray resonant transition radiation by few-MeV electrons traversing solid multilayer structures for the energies of interest to medicine (30-50 keV) changes dramatically compared with that for soft x-ray radiation. We show that one of the main factors in achieving the required resonant line is the absence of the contrast of the refractive indices between the spacer and the radiator at the far wings of the radiation line; for that purpose, the optimal spacer, as a rule, should have a higher atomic number than the radiator. Having experimental goals in mind, we have considered also the unwanted effects due to bremsstrahlung radiation, absorption and scattering of radiated photons, detector-related issues, and inhibited coherence of transition radiation due to random deviation of spacing between the layers. Choosing as a model example a Mo-Ag radiator-spacer pair of materials, we demonstrate that the x-ray transition radiation line can be well resolved with the use of spatial and frequency filtering

  4. Method for determining the irradiation dose deposited in a scintillator by ionising radiation and associated device - WO 2013060745 A1

    International Nuclear Information System (INIS)

    2013-01-01

    The invention relates to a method for determining an irradiation dose deposited in a scintillator (5) by ionising radiation, said method comprising the steps of: irradiating the scintillator (5) for a pre-determined time; detecting an instant at which the scintillator (5) is excited, using a first photodetector (11); subsequently, detecting an instant at which a scintillation photon is received, using a second photodetector (14) operating in single photon counting mode; identifying each sequence formed by the detection of an excitation instant by the first photodetector (11) and the detection of a reception instant by the second photodetector (14) at a coincidence event; counting the number of coincidence events; and obtaining the irradiation dose deposited during the irradiation time as a function of the number of coincidence events counted and a pre-determined proportionality factor. (authors)

  5. A more rugged ZnS(Ag) alpha scintillation detector

    International Nuclear Information System (INIS)

    McElhaney, S.A.; Ramsey, J.A.; Bauer, M.L.; Chiles, M.M.

    1990-01-01

    Conventional alpha scintillation detectors comprise a phosphor-coated light-pipe covered by a thin aluminized Mylar layer. This opaque radiation entrance window serves as a shield against ambient light entering the detector with minimum alpha attenuation. Unfortunately, Mylar is extremely fragile and easily punctured or torn by sticks, stones, and screws encountered during regular radiation surveys. The authors have been developing an alpha scintillation detector more rugged and durable than conventional models. This paper presents the scintillator assembly, which consists of a mixture of silver-activated zinc sulfide [ZnS(Ag)] and clear epoxy. The ZnS(Ag) scintillation powder is mixed with a low-viscosity, optically transparent epoxy and poured into a glass-smooth mold of desired shape and size

  6. The use of energy information in plastic scintillator materia

    International Nuclear Information System (INIS)

    Ely, J.; Anderson, K.; Bates, D.; Kouzes, R.; Lo Presti, C.; Runkle, R.; Siciliano, E.; Weier, D.

    2008-01-01

    Plastic scintillator material is often used for gamma-ray detection in many applications due to its relatively good sensitivity and cost-effectiveness compared to other detection materials. However, due to the dominant Compton scattering interaction mechanism, full energy peaks are not observed in plastic scintillator spectra and isotopic identification is impossible. Typically plastic scintillator detectors are solely gross count detectors. In some safeguards and security applications, such as radiation portal monitors for vehicle screening, naturally-occurring radioactive material (NORM) often triggers radiation alarms and results in innocent or nuisance alarms. The limited energy information from plastic scintillator material can be used to discriminate the NORM from targeted materials and reduce the nuisance alarm rate. An overview of the utilization of the energy information from plastic scintillator material will be presented, with emphasis on the detection capabilities and potential limitations for safeguards and security applications. (author)

  7. The Origins of Scintillator Non-Proportionality

    Science.gov (United States)

    Moses, W. W.; Bizarri, G. A.; Williams, R. T.; Payne, S. A.; Vasil'ev, A. N.; Singh, J.; Li, Q.; Grim, J. Q.; Choong, W.-S.

    2012-10-01

    Recent years have seen significant advances in both theoretically understanding and mathematically modeling the underlying causes of scintillator non-proportionality. The core cause is that the interaction of radiation with matter invariably leads to a non-uniform ionization density in the scintillator, coupled with the fact that the light yield depends on the ionization density. The mechanisms that lead to the luminescence dependence on ionization density are incompletely understood, but several important features have been identified, notably Auger-like processes (where two carriers of excitation interact with each other, causing one to de-excite non-radiatively), the inability of excitation carriers to recombine (caused either by trapping or physical separation), and the carrier mobility. This paper reviews the present understanding of the fundamental origins of scintillator non-proportionality, specifically the various theories that have been used to explain non-proportionality.

  8. Alkali metal hafnium oxide scintillators

    Science.gov (United States)

    Bourret-Courchesne, Edith; Derenzo, Stephen E.; Taylor, Scott Edward

    2018-05-08

    The present invention provides for a composition comprising an inorganic scintillator comprising an alkali metal hafnate, optionally cerium-doped, having the formula A2HfO3:Ce; wherein A is an alkali metal having a valence of 1, such as Li or Na; and the molar percent of cerium is 0% to 100%. The alkali metal hafnate are scintillators and produce a bright luminescence upon irradiation by a suitable radiation.

  9. Influence of the reactor irradiation on the radiation-optical features of the PbWO4:La scintillation crystals

    International Nuclear Information System (INIS)

    Ashurov, M.Kh.; Ismoilov, Sh.Kh.; Khatamov, K.; Gasanov, Eh.M.; Rustamov, I.R.

    2001-01-01

    Within an International LHC project the lead tungstates (PbWO 4 ) scintillation crystals radiation stability activated by La ions was carried out. In the 400-700 nm length range the transmission spectra were measured on the different parts of the standard PbWO 4 :La crystals. The spectra were measured before and after irradiation by both fast neutrons and γ-radiation. On the base of obtained data the contribution of γ-quanta and neutrons in the radiation-induced losses value of optical radiation in the active media of the electromagnetic colorimeter was estimated

  10. Plastic scintillators with β-diketone Eu complexes for high ionizing radiation detection

    International Nuclear Information System (INIS)

    Adadurov, A.F.; Zhmurin, P.N.; Lebedev, V.N.; Kovalenko, V.N.

    2011-01-01

    Luminescent and scintillation properties of polystyrene-based plastic scintillators with β-diketone Eu complexes are investigated. A scintillator with dibenzoylmethane Eu complex containing two phenyl groups demonstrates the maximum scintillating efficiency. It is shown that plastic scintillators efficiency is dramatically decreased if β-diketone derivatives contain no phenyl groups as substituents. This fact can be explained by exciplex mechanism of energy transfer from a matrix to Eu complex. - Highlights: → Fluorescent properties of polystyrene scintillators with β-diketone complexes of Eu were studied. → Scintillating efficiency is increased with the number of phenyl groups in Eu complex. → This is related to exciplex mechanism of energy transfer from a polymer matrix to Eu complex.

  11. Microfluidic Scintillation Detectors

    CERN Multimedia

    Microfluidic scintillation detectors are devices of recent introduction for the detection of high energy particles, developed within the EP-DT group at CERN. Most of the interest for such technology comes from the use of liquid scintillators, which entails the possibility of changing the active material in the detector, leading to an increased radiation resistance. This feature, together with the high spatial resolution and low thickness deriving from the microfabrication techniques used to manufacture such devices, is desirable not only in instrumentation for high energy physics experiments but also in medical detectors such as beam monitors for hadron therapy.

  12. Collimator trans-axial tomographic scintillation camera

    International Nuclear Information System (INIS)

    Jaszczak, Ronald J.

    1979-01-01

    An improved collimator is provided for a scintillation camera system that employs a detector head for transaxial tomographic scanning. One object of this invention is to significantly reduce the time required to obtain statistically significant data in radioisotope scanning using a scintillation camera. Another is to increase the rate of acceptance of radioactive events to contribute to the positional information obtainable from a radiation source of known strength without sacrificing spatial resolution. A further object is to reduce the necessary scanning time without degrading the images obtained. The collimator described has apertures defined by septa of different radiation transparency. The septa are aligned to provide greater radiation shielding from gamma radiation travelling within planes perpendicular to the cranial-caudal axis and less radiation shielding from gamma radiation travelling within other planes. Septa may also define apertures such that the collimator provides high spatial resolution of gamma rays traveling within planes perpendicular to the cranial-caudal axis and directed at the detector and high radiation sensitivity to gamma radiation travelling other planes and indicated at the detector. (LL)

  13. Electron traps and scintillation mechanism in YAlO3:Ce and LuAlO3:Ce scintillators

    International Nuclear Information System (INIS)

    Wojtowicz, A.J.; Glodo, J.; Drozdowski, W.; Przegietka, K.R.

    1998-01-01

    In this paper we present the results of thermoluminescence, isothermal decay and scintillation light yield measurements on two isostructural scintillator materials, YAlO 3 :Ce and LuAlO 3 :Ce. In addition to the variety of deep traps identified by thermoluminescence and isothermal decays, scintillation light yield experiments demonstrate the presence in both materials of a number of relatively shallow traps. While the deep traps may reduce the scintillation light yield, they do not influence the kinetics of the process. The shallow traps, on the other hand, by interfering with the process of radiative recombination of charge carriers via Ce 3+ ions, can strongly affect not only the yield of the scintillation process but its kinetics as well. The presence of shallow traps provides a consistent explanation for a number of poorly understood relationships between the two scintillator materials, including a higher room temperature scintillation light yield and longer scintillation decay time in YAlO 3 :Ce, and a longer scintillation rise time in LuAlO 3 :Ce. Theoretical analysis indicates that elimination of these traps would make the two materials nearly identical in scintillator performance. Although the specific identity of all traps remains elusive, the performance of both scintillator materials is now, in practical terms, fully understood. (Copyright (c) 1998 Elsevier Science B.V., Amsterdam. All rights reserved.)

  14. The transient radiation effects and hardness of programmed device

    International Nuclear Information System (INIS)

    Du Chuanhua; Xu Xianguo; Zhao Hailin

    2014-01-01

    A review and summary of research and development in the investigation of transient ionizing radiation effects in device and cirviut is presented. The transient ionizing radiation effects in two type of programmed device, that's 32 bit Microcontroller and antifuse FPGA, were studied. The expeiment test data indicate: The transient ionizing radiation effects of 32 bit Microcontroller manifested self-motion restart and Latchup, the Latchup threshold was 5 × 10"7 Gy (Si)/s. The transient ionizing radiation effects of FPGA was reset, no Latchup. The relationship of circuit effects to physical mechanisms was analized. A new method of hardness in circiut design was put forward. (authors)

  15. Discrimination of Naturally Occurring Radioactive Material in Plastic Scintillator Material

    International Nuclear Information System (INIS)

    Ely, James H.; Kouzes, Richard T.; Geelhood, Bruce D.; Schweppe, John E.; Warner, Ray A.

    2003-01-01

    Plastic scintillator material is used in many applications for the detection of gamma-rays from radioactive material, primarily due to the sensitivity per unit cost compared to other detection materials. However, the resolution and lack of full-energy peaks in the plastic scintillator material prohibits detailed spectroscopy. Therefore, other materials such as doped sodium iodide are used for spectroscopic applications. The limited spectroscopic information can however be exploited in plastic scintillator materials to provide some discrimination. The discrimination between man-made and naturally occurring sources would be useful in reducing alarm screening for radiation detection applications which target man-made sources. The results of applying the limited energy information from plastic scintillator material for radiation portal monitors are discussed.

  16. Room Temperature Hard Radiation Detectors Based on Solid State Compound Semiconductors: An Overview

    Science.gov (United States)

    Mirzaei, Ali; Huh, Jeung-Soo; Kim, Sang Sub; Kim, Hyoun Woo

    2018-05-01

    Si and Ge single crystals are the most common semiconductor radiation detectors. However, they need to work at cryogenic temperatures to decrease their noise levels. In contrast, compound semiconductors can be operated at room temperature due to their ability to grow compound materials with tunable densities, band gaps and atomic numbers. Highly efficient room temperature hard radiation detectors can be utilized in biomedical diagnostics, nuclear safety and homeland security applications. In this review, we discuss room temperature compound semiconductors. Since the field of radiation detection is broad and a discussion of all compound materials for radiation sensing is impossible, we discuss the most important materials for the detection of hard radiation with a focus on binary heavy metal semiconductors and ternary and quaternary chalcogenide compounds.

  17. Room Temperature Hard Radiation Detectors Based on Solid State Compound Semiconductors: An Overview

    Science.gov (United States)

    Mirzaei, Ali; Huh, Jeung-Soo; Kim, Sang Sub; Kim, Hyoun Woo

    2018-03-01

    Si and Ge single crystals are the most common semiconductor radiation detectors. However, they need to work at cryogenic temperatures to decrease their noise levels. In contrast, compound semiconductors can be operated at room temperature due to their ability to grow compound materials with tunable densities, band gaps and atomic numbers. Highly efficient room temperature hard radiation detectors can be utilized in biomedical diagnostics, nuclear safety and homeland security applications. In this review, we discuss room temperature compound semiconductors. Since the field of radiation detection is broad and a discussion of all compound materials for radiation sensing is impossible, we discuss the most important materials for the detection of hard radiation with a focus on binary heavy metal semiconductors and ternary and quaternary chalcogenide compounds.

  18. 2003: A centennial of spinthariscope and scintillation counting

    International Nuclear Information System (INIS)

    Kolar, Z.I.; Hollander, W. den

    2004-01-01

    In 1903 W. Crookes demonstrated in England his 'spinthariscope' for the visual observation of individual scintillations caused by alpha particles impinging upon a ZnS screen. In contrast to the analogue methods of radiation measurements in that time the spinthariscope was a single-particle counter, being the precursor of scintillation counters since. In the same period F. Giesel, J. Elster and H. Geitel in Germany also found that scintillations from ZnS represent single particle events. This paper summarises the historical events relevant to the advent of scintillation counting

  19. Performance And Radiation Hardness Of The Atlas/sct Detector Module

    CERN Document Server

    Eklund, L

    2003-01-01

    The ATLAS experiment is a general purpose experiment being constructed at the Large Hadron Collider (LHC) at FERN, Geneva. ATLAS is designed to exploit the full physics potential of LHC, in particular to study topics concerning the Higgs mechanism, Super-symmetry and CP violation. The cross sections for the processes under study are extremely small, requiring very high luminosity colliding beams. The Semiconductor Tracker (SCT) is an essential part of the Inner Detector tracking system of ATLAS. The active elements of the SCT is 4088 detector modules, tiled on four barrel cylinders and eighteen endcap disks. As a consequence of the high luminosity, the detector modules will operate in a harsh radiation environment. This thesis describes work concerning radiation hardness, beam test performance and methods for production testing of detector modules. The radiation hardness studies have been focused on the electrical performance of the front-end ASIC and the detector module. The results have identified features ...

  20. Investigation of radiation absorption and X-ray fluorescence properties of medical imaging scintillators by Monte Carlo methods

    International Nuclear Information System (INIS)

    Nikolopoulos, D.; Kandarakis, I.; Cavouras, D.; Valais, I.; Linardatos, D.; Michail, C.; David, S.; Gaitanis, A.; Nomicos, C.; Louizi, A.

    2006-01-01

    X-ray absorption and X-ray fluorescence properties of medical imaging scintillating screens were studied by Monte Carlo methods as a function of the incident photon energy and screen-coating thickness. The scintillating materials examined were Gd 2 O 2 S (GOS) Gd 2 SiO 5 (GSO) YAlO 3 (YAP), Y 3 Al 5 O 12 (YAG), LuSiO 5 (LSO), LuAlO 3 (LuAP) and ZnS. Monoenergetic photon exposures were modeled in the range from 10 to 100 keV. The corresponding ranges of coating thicknesses of the investigated scintillating screens ranged up to 200 mg cm -2 . Results indicated that X-ray absorption and X-ray fluorescence are affected by the incident photon energy and the screen's coating thickness. Regarding incident photon energy, this X-ray absorption and fluorescence was found to exhibit very intense changes near the corresponding K edge of the heaviest element in the screen's scintillating material. Regarding coating thickness, thicker screens exhibited higher X-ray absorption and X-ray fluorescence. Results also indicated that a significant fraction of the generated X-ray fluorescent quanta escape from the scintillating screen. This fraction was found to increase with screen's coating thickness. At the energy range studied, most of the incident photons were found to be absorbed via one-hit photoelectric effect. As a result, the reabsorption of scattered radiation was found to be of rather minor importance; nevertheless this was found to increase with the screen's coating thickness. Differences in X-ray absorption and X-ray fluorescence were found among the various scintillators studied. LSO scintillator was found to be the most attractive material for use in many X-ray imaging applications, exhibiting the best absorption properties in the largest part of the energy range studied. Y-based scintillators were also found to be of significant absorption performance within the low energy ranges

  1. Use of GaN as a Scintillating Ionizing Radiation Detector

    Science.gov (United States)

    Wensman, Johnathan; Guardala, Noel; Mathur, Veerendra; Alasagas, Leslie; Vanhoy, Jeffrey; Statham, John; Marron, Daniel; Millett, Marshall; Marsh, Jarrod; Currie, John; Price, Jack

    2017-09-01

    Gallium nitride (GaN) is a III/V direct bandgap semiconductor which has been used in light emitting diodes (LEDs) since the 1990s. Currently, due to a potential for increased efficiency, GaN is being investigated as a replacement for silicon in power electronics finding potential uses ranging from data centers to electric vehicles. In addition to LEDs and power electronics though, doped GaN can be used as a gamma insensitive fast neutron detector due to the direct band-gap, light propagation properties, and response to ionizing radiations. Investigation of GaN as a semiconductor scintillator for use in a radiation detection system involves mapping the response function of the detector crystal over a range of photon and neutron energies, and measurements of light generation in the GaN crystal due to proton, alpha, and nitrogen projectiles. In this presentation we discuss the measurements made to date, and plausible interpretations of the response functions. This work funded in part by the Naval Surface Warfare Center, Carderock Division In-house Laboratory Independent Research program.

  2. GaN-Based High Temperature and Radiation-Hard Electronics for Harsh Environments

    Science.gov (United States)

    Son, Kyung-ah; Liao, Anna; Lung, Gerald; Gallegos, Manuel; Hatakeh, Toshiro; Harris, Richard D.; Scheick, Leif Z.; Smythe, William D.

    2010-01-01

    We develop novel GaN-based high temperature and radiation-hard electronics to realize data acquisition electronics and transmitters suitable for operations in harsh planetary environments. In this paper, we discuss our research on metal-oxide-semiconductor (MOS) transistors that are targeted for 500 (sup o)C operation and >2 Mrad radiation hardness. For the target device performance, we develop Schottky-free AlGaN/GaN MOS transistors, where a gate electrode is processed in a MOS layout using an Al2O3 gate dielectric layer....

  3. RD50 Collaboration overview: Development of new radiation hard detectors

    Energy Technology Data Exchange (ETDEWEB)

    Kuehn, S., E-mail: susanne.kuehn@cern.ch

    2016-07-11

    Silicon sensors are widely used as tracking detectors in high energy physics experiments. This results in several specific requirements like radiation hardness and granularity. Therefore research for highly performing silicon detectors is required. The RD50 Collaboration is a CERN R&D collaboration dedicated to the development of radiation hard silicon devices for application in high luminosity collider experiments. Extensive research is ongoing in different fields since 2001. The collaboration investigates both defect and material characterization, detector characterization, the development of new structures and full detector systems. The report gives selected results of the collaboration and places an emphasis on the development of new structures, namely 3D devices, CMOS sensors in HV technology and low gain avalanche detectors. - Highlights: • The RD50 Collaboration is a CERN R&D collaboration dedicated to the development of radiation hard silicon devices for high luminosity collider experiments. • The collaboration investigates defect, material and detector characterization, the development of new structures and full detector systems. • Results of measured data of n-in-p type sensors allow recommendations for silicon tracking detectors at the HL-LHC. • The charge multiplication effect was investigated to allow its exploitation and resulted in new structures like LGAD sensors. • New sensor types like slim and active edge sensors, 3D detectors, and lately HVCMOS devices were developed in the active collaboration.

  4. Study of hard braking x-ray radiation on the radiation-beam complex ''TEMP''

    International Nuclear Information System (INIS)

    Batrakov, A.B.; Glushko, E.G.; Egorov, A.M.; Zinchenko, A.A.; Litvinenko, V.V.; Lonin, Yu.F.; Ponomarev, A.G.; Rybka, A.V.; Fedotov, S.I.; Uvarov, V.T.

    2015-01-01

    A calculation over of basic parameters of the hard brake x-rayed radiation for the microsecond accelerating of relativistic electronic beam T EMP . Optimization of converters is conducted for these aims. Maximal doses are experimentally got brake x-rayed radiation on beam-radiation complex T EMP . The diagrams of orientation of the brake x-rayed radiation are taken off depending on energies of bunches and forms of electrodes.

  5. Development and radiation evaluation of mobile station for personnel monitoring system based on indigenous plastic scintillator sensor rods

    International Nuclear Information System (INIS)

    Chaudhary, H.S.; Parihar, A.; Senwar, K.R.; Prakash, V.; Rathore, A.S.

    2018-01-01

    The Mobile Station for Personnel Monitoring (MSPM) system has been designed and developed for rapid screening of personnel with respect to radiation contamination during nuclear or radiological emergency; it can also be used for prevention of illicit movement of radioactive sources. The objective was to develop a modular, transportable and easily deployable gamma portal monitoring system based on indigenous DLJ developed plastic scintillator sensors. The Gamma radiation response of the system is presented here

  6. Recent results on the development of radiation-hard diamond detectors

    CERN Document Server

    Conway, J S; Bauer, C; Berdermann, E; Bergonzo, P; Bogani, F; Borchi, E; Brambilla, A; Bruzzi, Mara; Colledani, C; Dabrowski, W; Da Graca, J; Delpierre, P A; Deneuville, A; Dulinski, W; van Eijk, B; Fallou, A; Fizzotti, F; Foulon, F; Friedl, M; Gan, K K; Gheeraert, E; Grigoriev, E; Hallewell, G D; Hall-Wilton, R; Han, S; Hartjes, F G; Hrubec, Josef; Husson, D; Jamieson, D; Kagan, H; Kania, D R; Kaplon, J; Karl, C; Kass, R; Knöpfle, K T; Krammer, Manfred; Lo Giudice, A; Lü, R; Manfredi, P F; Manfredotti, C; Marshall, R D; Meier, D; Mishina, M; Oh, A; Pan, L S; Palmieri, V G; Pernicka, Manfred; Peitz, A; Pirollo, S; Plano, R; Polesello, P; Prawer, S; Pretzl, Klaus P; Procario, M; Re, V; Riester, J L; Roe, S; Roff, D G; Rudge, A; Russ, J; Schnetzer, S; Sciortino, S; Somalwar, S V; Speziali, V; Stelzer, H; Stone, R; Suter, B; Tapper, R J; Tesarek, R; Thomson, G B; Trawick, M; Trischuk, W; Vittone, E; Walsh, A M; Wedenig, R; Weilhammer, Peter; White, C; Ziock, H J; Zöller, M

    1999-01-01

    Charged particle detectors made from chemical vapor deposition (CVD) diamond have radiation hardness greatly exceeding that of silicon- based detectors. The CERN-based RD42 Collaboration has developed and tested CVD diamond microstrip and pixel detectors with an eye to their application in the intense radiation environment near the interaction region of hadron colliders. This paper presents recent results from tests of these detectors. (4 refs).

  7. Radiation imaging with a new scintillator and a CMOS camera

    Czech Academy of Sciences Publication Activity Database

    Kurosawa, S.; Shoji, Y.; Pejchal, Jan; Yokota, Y.; Yoshikawa, A.

    2015-01-01

    Roč. 9, Jul (2015), C07015 ISSN 1748-0221 Institutional support: RVO:68378271 Keywords : scintillators * scintillation and light emission processes * image processin Subject RIV: BL - Plasma and Gas Discharge Physics Impact factor: 1.310, year: 2015

  8. Scintillator plate calorimetry

    International Nuclear Information System (INIS)

    Price, L.E.

    1990-01-01

    Calorimetry using scintillator plates or tiles alternated with sheets of (usually heavy) passive absorber has been proven over multiple generations of collider detectors. Recent detectors including UA1, CDF, and ZEUS have shown good results from such calorimeters. The advantages offered by scintillator calorimetry for the SSC environment, in particular, are speed (<10 nsec), excellent energy resolution, low noise, and ease of achieving compensation and hence linearity. On the negative side of the ledger can be placed the historical sensitivity of plastic scintillators to radiation damage, the possibility of nonuniform response because of light attenuation, and the presence of cracks for light collection via wavelength shifting plastic (traditionally in sheet form). This approach to calorimetry is being investigated for SSC use by a collaboration of Ames Laboratory/Iowa State University, Argonne National Laboratory, Bicron Corporation, Florida State University, Louisiana State University, University of Mississippi, Oak Ridge National Laboratory, Virginia Polytechnic Institute and State University, Westinghouse Electric Corporation, and University of Wisconsin

  9. CdTe Timepix detectors for single-photon spectroscopy and linear polarimetry of high-flux hard x-ray radiation.

    Science.gov (United States)

    Hahn, C; Weber, G; Märtin, R; Höfer, S; Kämpfer, T; Stöhlker, Th

    2016-04-01

    Single-photon spectroscopy of pulsed, high-intensity sources of hard X-rays - such as laser-generated plasmas - is often hampered by the pileup of several photons absorbed by the unsegmented, large-volume sensors routinely used for the detection of high-energy radiation. Detectors based on the Timepix chip, with a segmentation pitch of 55 μm and the possibility to be equipped with high-Z sensor chips, constitute an attractive alternative to commonly used passive solutions such as image plates. In this report, we present energy calibration and characterization measurements of such devices. The achievable energy resolution is comparable to that of scintillators for γ spectroscopy. Moreover, we also introduce a simple two-detector Compton polarimeter setup with a polarimeter quality of (98 ± 1)%. Finally, a proof-of-principle polarimetry experiment is discussed, where we studied the linear polarization of bremsstrahlung emitted by a laser-driven plasma and found an indication of the X-ray polarization direction depending on the polarization state of the incident laser pulse.

  10. Shock-resistant scintillation detector

    International Nuclear Information System (INIS)

    Novak, W.P.

    1979-01-01

    A unique scintillation detector unit is disclosed which employs a special light transfer and reflector means that encases and protects the scintillator crystal against high g forces. The light transfer means comprises a flexible silicon rubber optical material bonded between the crystal and the optical window and having an axial thickness sufficient to allow the scintillator to move axially inside the container under high g forces without destroying the bonds. The reflector means comprises a soft elastic silicone rubber sleeve having a multiplicity of closely arranged tapered protrusions radiating toward and engaging the periphery of the scintillator crystal to cushion shocks effectively and having a reflective material, such as aluminum oxide powder, in the spaces between the protrusions. The reflector means provides improved shock absorption because of the uniform support and cushioning action of the protrusions and also provides the detector with high efficiency. The silicon rubber composition is specially compounded to include a large amount of aluminum oxide which enables the rubber to function effectively as a light reflector

  11. Monte Carlo modeling of fiber-scintillator flow-cell radiation detector geometry

    International Nuclear Information System (INIS)

    Rucker, T.L.; Ross, H.H.; Tennessee Univ., Knoxville; Schweitzer, G.K.

    1988-01-01

    A Monte Carlo computer calculation is described which models the geometric efficiency of a fiber-scintillator flow-cell radiation detector designed to detect radiolabeled compounds in liquid chromatography eluates. By using special mathematical techniques, an efficiency prediction with a precision of 1% is obtained after generating only 1000 random events. Good agreement is seen between predicted and experimental efficiency except for very low energy beta emission where the geometric limitation on efficiency is overcome by pulse height limitations which the model does not consider. The modeling results show that in the test system, the detection efficiency for low energy beta emitters is limited primarily by light generation and collection rather than geometry. (orig.)

  12. Hardness measurements of silicon rubber and polyurethane rubber cured by ionizing radiation

    International Nuclear Information System (INIS)

    Basfar, A.A.

    1995-01-01

    This work investigates the hardness of both silicon rubber and polyurethane rubber cured by ionizing radiation. Shore A Hardness is used to characterize the subject elastomers in relation to the crosslinking process. Various formulations of both materials have been investigated in order to achieve the optimum cure conditions desired. A small amount of the curing agent has been incorporated in some formulations in order to reduce the required dose to achieve full cure conditions. Silicon rubber has shown improvements in hardness as absorbed dose is increased, whereas hardness remained constant over a range of absorbed doses for polyurethane rubber

  13. Development of crystals based in cesium iodide for application as radiation detectors

    International Nuclear Information System (INIS)

    Pereira, Maria da Conceicao Costa

    2006-01-01

    Inorganic scintillators with fast luminescence decay time, high density and high light output have been the object of studies for application in nuclear physics, high energy physics, nuclear tomography and other fields of science and engineering. Scintillation crystals based on cesium iodide (CsI) are matters with relatively low higroscopy, high atomic number, easy handling and low cost, characteristics that favor their use as radiation detectors. In this work, the growth of pure CsI crystals, CsI:Br and CsI:Pb, using the Bridgman technique, is described. The concentration of the bromine doping element (Br) was studied in the range of 1,5x10 -1 M to 10 -2 M and the lead (Pb) in the range of 10 -2 M to 5x10 -4 M. To evaluate the scintillators developed, systematic measurements were carried out for luminescence emission and luminescence decay time for gamma radiation, optical transmittance assays, Vickers micro-hardness assays, determination of the doping elements distribution along the grown crystals and analysis of crystals response to the gamma radiation in the energy range of 350 keV to 1330 keV and alpha particles from a 241 Am source, with energy of 5.54 MeV. It was obtained 13 ns to 19 ns for luminescence decay time for CsI:Br and CsI:Pb crystals. These results were very promising. The results obtained for micro-hardness showed a significant increase in function of the doping elements concentration, when compared to the pure CsI crystal, increasing consequently the mechanical resistance of the grown crystals. The validity of using these crystals as radiation sensors may be seen from the results of their response to gamma radiation and alpha particles. (author)

  14. Scintillating plastic optical fiber radiation detectors in high energy particle physics

    International Nuclear Information System (INIS)

    Bross, A.D.

    1991-01-01

    We describe the application of scintillating optical fiber in instrumentation for high energy particle physics. The basic physics of the scintillation process in polymers is discussed first and then we outline the fundamentals of scintillating fiber technology. Fiber performance, optimization, and characterization measurements are given. Detector applications in the areas of particle tracking and particle energy determination are then described. 13 refs., 12 figs

  15. Inclusion of Radiation Environment Variability in Total Dose Hardness Assurance Methodology

    Science.gov (United States)

    Xapsos, M. A.; Stauffer, C.; Phan, A.; McClure, S. S.; Ladbury, R. L.; Pellish, J. A.; Campola, M. J.; LaBel, K. A.

    2016-01-01

    Variability of the space radiation environment is investigated with regard to parts categorization for total dose hardness assurance methods. It is shown that it can have a significant impact. A modified approach is developed that uses current environment models more consistently and replaces the radiation design margin concept with one of failure probability during a mission.

  16. Proportional gas scintillation detectors and their applications

    International Nuclear Information System (INIS)

    Petr, I.

    1978-01-01

    The principle is described of a gas proportional scintillation detector and its function. Dependence of Si(Li) and xenon proportional detectors energy resolution on the input window size is given. A typical design is shown of a xenon detector used for X-ray spetrometry at an energy of 277 eV to 5.898 keV and at a gas pressure of 98 to 270 kPa. Gas proportional scintillation detectors show considerable better energy resolution than common proportional counters and even better resolution than semiconductor Si(Li) detectors for low X radiation energies. For detection areas smaller than 25 mm 2 Si(Li) detectors show better resolution, especially for higher X radiation energies. For window areas 25 to 190 mm 2 both types of detectors are equal, for a window area exceeding 190 mm 2 the proportional scintillation detector has higher energy resolution. (B.S.)

  17. Barium iodide and strontium iodide crystals andd scintillators implementing the same

    Science.gov (United States)

    Payne, Stephen A; Cherepy, Nerine J; Hull, Giulia E; Drobshoff, Alexander D; Burger, Arnold

    2013-11-12

    In one embodiment, a material comprises a crystal comprising strontium iodide providing at least 50,000 photons per MeV. A scintillator radiation detector according to another embodiment includes a scintillator optic comprising europium-doped strontium iodide providing at least 50,000 photons per MeV. A scintillator radiation detector in yet another embodiment includes a scintillator optic comprising SrI.sub.2 and BaI.sub.2, wherein a ratio of SrI.sub.2 to BaI.sub.2 is in a range of between 0:1 A method for manufacturing a crystal suitable for use in a scintillator includes mixing strontium iodide-containing crystals with a source of Eu.sup.2+, heating the mixture above a melting point of the strontium iodide-containing crystals, and cooling the heated mixture near the seed crystal for growing a crystal. Additional materials, systems, and methods are presented.

  18. Physical chemical radiation effects in scintillator solutions content PPO and POPOP

    International Nuclear Information System (INIS)

    Fernandes Neto, Jose Mari; Hamada, Margarida Mizue; Duarte, Celina Lopes; Mesquita, Carlos Henrique de

    2005-01-01

    Samples containing PPO (1%, g/mL), diluted in toluene, were irradiated in a 60 Co irradiator (6.46 kGy/h) at different doses. The PPO concentration decay bi-exponentially with the dose, generating the degradation products: benzoic acid, benzamide and benzilic alcohol. The liquid scintillator system was not sensitive to the radiation damage until 20 kGy. Otherwise, the pulse height analysis showed that doses between 30 and 40 kGy generated significant loss of quality of the sensor (scintillating liquid) and the light yield was reduced in half with the dose of (34.04 ± 0.80) kGy. This value was confirmed by the photopeak position analysis that resulted in D1/2 = (31.7 ± 1.4) kGy. The transmittance, at 360 nm, of the irradiated solution decreased exponentially. The compartmental model using five compartments (fast decay PPO, slow decay PPO, benzamide, benzoic acid and benzilic alcohol) was satisfactory to explain the decay of the PPO in its degradation products as a function of the dose. The explanation coefficient r2 = 0.985636 demonstrates that the model was capable to explain 98.6% of the experimental variations. The energies involved in the chemical reactions were w = (0.239 ± 0.031) eV/damage (fast decay) and w (1.834 ± 0.301) eV/damage (slow decay). (author)

  19. BC404 scintillators as gamma locators studied via Geant4 simulations

    Science.gov (United States)

    Cortés, M. L.; Hoischen, R.; Eisenhauer, K.; Gerl, J.; Pietralla, N.

    2014-05-01

    In many applications in industry and academia, an accurate determination of the direction from where gamma rays are emitted is either needed or desirable. Ion-beam therapy treatments, the search for orphan sources, and homeland security applications are examples of fields that can benefit from directional sensitivity to gamma-radiation. Scintillation detectors are a good option for these types of applications as they have relatively low cost, are easy to handle and can be produced in a large range of different sizes. In this work a Geant4 simulation was developed to study the directional sensitivity of different BC404 scintillator geometries and arrangements. The simulation includes all the physical processes relevant for gamma detection in a scintillator. In particular, the creation and propagation of optical photons inside the scintillator was included. A simplified photomultiplier tube model was also simulated. The physical principle exploited is the angular dependence of the shape of the energy spectrum obtained from thin scintillator layers when irradiated from different angles. After an experimental confirmation of the working principle of the device and a check of the simulation, the possibilities and limitations of directional sensitivity to gamma radiation using scintillator layers was tested. For this purpose, point-like sources of typical energies expected in ion-beam therapy were used. Optimal scintillator thicknesses for different energies were determined and the setup efficiencies calculated. The use of arrays of scintillators to reconstruct the direction of incoming gamma rays was also studied. For this case, a spherical source emitting Bremsstrahlung radiation was used together with a setup consisting of scintillator layers. The capability of this setup to identify the center of the extended source was studied together with its angular resolution.

  20. Liquid scintillation counting (LSC) - an overview

    International Nuclear Information System (INIS)

    Ravi, S.; Mathew, K.M.

    2005-01-01

    In Liquid Scintillation Counting, the amount of light produced is proportional to the amount of radiation present in the sample and the energy of the light produced is proportional to the energy of the radiation that is present in the sample. This makes LSC a very convenient tool to measure radioactivity

  1. LiCaAlF sub 6 :Ce crystal: a new scintillator

    CERN Document Server

    Gektin, A V; Neicheva, S; Gavrilyuk, V; Bensalah, A; Fukuda, T; Shimamura, K

    2002-01-01

    Scintillation properties of LiCaAlF sub 6 :Ce crystal, well known as the effective UV laser material, is reported. Ce sup 3 sup + emission at 286-305 nm with a single exponential decay time of 35 ns provides a scintillation pulse. Radiation damage in pure and Ce-doped crystals is studied. In contrast to the majority of fluoride crystals, cerium is responsible for the ultradeep traps formation revealing thermostimulated luminescence. Overlapping of color center absorption and Ce sup 3 sup + ion emission bands limits the scintillation efficiency of LiCaAlF sub 6 :Ce at high radiation doses.

  2. Intelligent optimization models based on hard-ridge penalty and RBF for forecasting global solar radiation

    International Nuclear Information System (INIS)

    Jiang, He; Dong, Yao; Wang, Jianzhou; Li, Yuqin

    2015-01-01

    Highlights: • CS-hard-ridge-RBF and DE-hard-ridge-RBF are proposed to forecast solar radiation. • Pearson and Apriori algorithm are used to analyze correlations between the data. • Hard-ridge penalty is added to reduce the number of nodes in the hidden layer. • CS algorithm and DE algorithm are used to determine the optimal parameters. • Proposed two models have higher forecasting accuracy than RBF and hard-ridge-RBF. - Abstract: Due to the scarcity of equipment and the high costs of maintenance, far fewer observations of solar radiation are made than observations of temperature, precipitation and other weather factors. Therefore, it is increasingly important to study several relevant meteorological factors to accurately forecast solar radiation. For this research, monthly average global solar radiation and 12 meteorological parameters from 1998 to 2010 at four sites in the United States were collected. Pearson correlation coefficients and Apriori association rules were successfully used to analyze correlations between the data, which provided a basis for these relative parameters as input variables. Two effective and innovative methods were developed to forecast monthly average global solar radiation by converting a RBF neural network into a multiple linear regression problem, adding a hard-ridge penalty to reduce the number of nodes in the hidden layer, and applying intelligent optimization algorithms, such as the cuckoo search algorithm (CS) and differential evolution (DE), to determine the optimal center and scale parameters. The experimental results show that the proposed models produce much more accurate forecasts than other models

  3. Scintillation camera with second order resolution

    International Nuclear Information System (INIS)

    Muehllehner, G.

    1976-01-01

    A scintillation camera for use in radioisotope imaging to determine the concentration of radionuclides in a two-dimensional area is described in which means is provided for second order positional resolution. The phototubes, which normally provide only a single order of resolution, are modified to provide second order positional resolution of radiation within an object positioned for viewing by the scintillation camera. The phototubes are modified in that multiple anodes are provided to receive signals from the photocathode in a manner such that each anode is particularly responsive to photoemissions from a limited portion of the photocathode. Resolution of radioactive events appearing as an output of this scintillation camera is thereby improved

  4. Scintillation camera with second order resolution

    International Nuclear Information System (INIS)

    1975-01-01

    A scintillation camera is described for use in radioisotope imaging to determine the concentration of radionuclides in a two-dimensional area in which means is provided for second-order positional resolution. The phototubes which normally provide only a single order of resolution, are modified to provide second-order positional resolution of radiation within an object positioned for viewing by the scintillation camera. The phototubes are modified in that multiple anodes are provided to receive signals from the photocathode in a manner such that each anode is particularly responsive to photoemissions from a limited portion of the photocathode. Resolution of radioactive events appearing as an output of this scintillation camera is thereby improved

  5. Development of water radiocontamination monitor using a plastic scintillator detector

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Madi Filho, T.; Hamada, M.M.

    1990-01-01

    An alpha, beta and gamma radiation water monitor was developed using a plastic scintillator detector with a sensitivity level of 15 bplastic scintillator detector with a sensitivity level of 15 Bq.L -1 and a counting efficiency of 25% for 131 I. It was proposed to be used in the radiation monitoring program of the research reactor swimming-pool of Sao Paulo. A simplified design and some properties of this monitor are presented. (author) [pt

  6. Efficiency and yield spectra of inorganic scintillates

    International Nuclear Information System (INIS)

    Rodnyi, P.A.

    1998-01-01

    Recent developments in the field of energy loss in inorganic scintillators are reviewed. The main parameters, which control the fundamental limit of the scintillator energy efficiency, are determined. It is shown that together with simple cascade processes one should take into account the production of plasmons to estimate the energy efficiency of scintillators or other phosphors excited by an ionizing radiation. Core-to-valence luminescence related to 5pCs→3pCl transitions is investigated in some chlorides: CsCl, KCl, RbCl, NaCl, KCaCl 3 , RbCaCl 3 . The yield spectra of the crystals in the VUV and X-ray regions are also studied. It is shown that the 4pRb-core states are involved in the process of creation of holes in the 5pCs-core band in Rb-based crystals. The formation of holes in the potassium core band acts as a competing process and suppresses the radiative core-to-valence transitions

  7. Real-time volumetric scintillation dosimetry

    International Nuclear Information System (INIS)

    Beddar, S

    2015-01-01

    The goal of this brief review is to review the current status of real-time 3D scintillation dosimetry and what has been done so far in this area. The basic concept is to use a large volume of a scintillator material (liquid or solid) to measure or image the dose distributions from external radiation therapy (RT) beams in three dimensions. In this configuration, the scintillator material fulfills the dual role of being the detector and the phantom material in which the measurements are being performed. In this case, dose perturbations caused by the introduction of a detector within a phantom will not be at issue. All the detector configurations that have been conceived to date used a Charge-Coupled Device (CCD) camera to measure the light produced within the scintillator. In order to accurately measure the scintillation light, one must correct for various optical artefacts that arise as the light propagates from the scintillating centers through the optical chain to the CCD chip. Quenching, defined in its simplest form as a nonlinear response to high-linear energy transfer (LET) charged particles, is one of the disadvantages when such systems are used to measure the absorbed dose from high-LET particles such protons. However, correction methods that restore the linear dose response through the whole proton range have been proven to be effective for both liquid and plastic scintillators. Volumetric scintillation dosimetry has the potential to provide fast, high-resolution and accurate 3D imaging of RT dose distributions. Further research is warranted to optimize the necessary image reconstruction methods and optical corrections needed to achieve its full potential

  8. Set of counts by scintillations for atmospheric samplings

    International Nuclear Information System (INIS)

    Appriou, D.; Doury, A.

    1962-01-01

    The author reports the development of a scintillation-based counting assembly with the following characteristics: a photo-multiplier with a wide photo-cathode, a thin plastic scintillator for the counting of beta + alpha (and possibility of mounting an alpha scintillator), a relatively small own motion with respect to activities to be counted, a weakly varying efficiency. The authors discuss the counting objective, present equipment tests (counter, proportional amplifier and pre-amplifier, input drawer). They describe the apparatus operation, discuss the selection of scintillators, report the study of the own movement (electron-based background noise, total background noise, background noise reduction), discuss counts (influence of the external source, sensitivity to alpha radiations, counting homogeneity, minimum detectable activity) and efficiencies

  9. Scintillation properties of semiconducting {sup 6}LiInSe{sub 2} crystals to ionizing radiation

    Energy Technology Data Exchange (ETDEWEB)

    Wiggins, Brenden [Y-12 National Security Complex, Oak Ridge, TN (United States); Vanderbilt University, Nashville, TN (United States); Groza, Michael; Tupitsyn, Eugene [Fisk University, Nashville, TN (United States); Lukosi, Eric [University of Tennessee, Knoxville, TN (United States); Stassun, Keivan; Burger, Arnold [Vanderbilt University, Nashville, TN (United States); Fisk University, Nashville, TN (United States); Stowe, Ashley [Y-12 National Security Complex, Oak Ridge, TN (United States); Vanderbilt University, Nashville, TN (United States); University of Tennessee, Knoxville, TN (United States)

    2015-11-21

    {sup 6}LiInSe{sub 2} has gained attention recently as a semiconducting thermal neutron detector. As presented herein, the chalcogenide compound semiconductor also detects incident neutrons via scintillation, making {sup 6}LiInSe{sub 2} the only lithium containing semiconductor to respond to neutrons via both detection mechanisms. Both yellow and red crystals, which appear in the literature, were investigated. Only the yellow crystal responded favorably to ionizing radiation, similar to the semiconducting operation utilizing electrodes. The obtained light yield for yellow crystals is 4400 photons/MeV, referenced to Bi{sub 4}Ge{sub 3}O{sub 12} (BGO).The estimated thermal neutron light yield was 21,000 photons/thermal neutron. The two measured decay time components were found to be 31±1 ns (49%) and 143±9 ns (51%).This crystal provides efficient, robust detection of neutrons via scintillation with respectable light yield and rapid response, enabling its use for a broad array of neutron detection applications.

  10. Comparison of plastic scintillating fibres and capillaries filled with liquid scintillator

    International Nuclear Information System (INIS)

    Cardini, A.; Cavasinni, V.; Girolamo, B. di; Flaminio, V.; Golovkin, S.V.; Gorin, A.M.; Kulichenko, A.V.; Kushnirenko, A.E.; Pyshev, A.I.; Manuilov, I.; Vasilchenko, V.G.

    1994-01-01

    A comparison is made between the light yield, attenuation length, time response and light propagation speed in plastic scintillating fibres (SCSF-38 and Kuraray-3HF) and quartz capillaries filled with liquid scintillator (LS) 1-methilnaphthalene (1MN) doped with new dyes R45 and R39. The inner diameter of capillaries and diameter of plastic fibres is 0.5 mm. The number of photoelectrons detected at the far end (2 m) was 2.9 for capillaries filled with 1MN+3 g/l R45 while it was 1.8 times smaller in the case of SCSF-38 and 3 times smaller in the case of Kuraray 3HF plastic fibres. Taking into account the quantum efficiency of the photodetector used these reduction factors became 3.0 and 2.0, respectively. Good attenuation length, high light output and also excellent radiation resistance of capillaries filled with LS (>60 Mrad, measured elsewhere) show that they are a very promising alternative to plastic scintillating fibres for future applications in tracking detectors and calorimeters. ((orig.))

  11. Developments of scintillator-based soft x-ray diagnostic in LHD with CsI:Tl and P47 scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Bando, T., E-mail: bando.takahiro@nifs.ac.jp [SOKENDAI (The Graduate University for Advanced Studies), 322-6 Oroshi-cho, Toki 509-5292 (Japan); Ohdachi, S.; Suzuki, Y. [SOKENDAI (The Graduate University for Advanced Studies), 322-6 Oroshi-cho, Toki 509-5292 (Japan); National Institute for Fusion Science, 322-6 Oroshi-cho, Toki 509-5292 (Japan)

    2016-11-15

    Multi-channel soft x-ray (SX) diagnostic has been used in the large helical device (LHD) to research magnetohydrodynamic equilibria and activities. However, in the coming deuterium plasma experiments of LHD, it will be difficult to use semiconductor systems near LHD. Therefore, a new type of SX diagnostic, a scintillator-based type diagnostic, has been investigated in order to avoid damage from the radiation. A fiber optic plate coated by P47 scintillator will be used to detect SX emission. Scintillation light will be transferred by pure silica core optical fibers and detected by photomultiplier tubes. A vertically elongated section of LHD will be covered by a 13 ch. array. Effects from the Deuterium Deuterium neutrons can be negligible when the scintillator is covered by a Pb plate 4 cm in thickness to avoid gamma-rays.

  12. Radiation-hard/high-speed parallel optical links

    Energy Technology Data Exchange (ETDEWEB)

    Gan, K.K., E-mail: gan@mps.ohio-state.edu [Department of Physics, The Ohio State University, Columbus, OH 43210 (United States); Buchholz, P.; Heidbrink, S. [Fachbereich Physik, Universität Siegen, Siegen (Germany); Kagan, H.P.; Kass, R.D.; Moore, J.; Smith, D.S. [Department of Physics, The Ohio State University, Columbus, OH 43210 (United States); Vogt, M.; Ziolkowski, M. [Fachbereich Physik, Universität Siegen, Siegen (Germany)

    2016-09-21

    We have designed and fabricated a compact parallel optical engine for transmitting data at 5 Gb/s. The device consists of a 4-channel ASIC driving a VCSEL (Vertical Cavity Surface Emitting Laser) array in an optical package. The ASIC is designed using only core transistors in a 65 nm CMOS process to enhance the radiation-hardness. The ASIC contains an 8-bit DAC to control the bias and modulation currents of the individual channels in the VCSEL array. The performance of the optical engine up at 5 Gb/s is satisfactory.

  13. WE-AB-BRB-03: Real-Time Volumetric Scintillation Dosimetry for Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Beddar, S. [UT MD Anderson Cancer Center (United States)

    2016-06-15

    Despite widespread IMRT treatments at modern radiation therapy clinics, precise dosimetric commissioning of an IMRT system remains a challenge. In the most recent report from the Radiological Physics Center (RPC), nearly 20% of institutions failed an end-to-end test with an anthropomorphic head and neck phantom, a test that has rather lenient dose difference and distance-to-agreement criteria of 7% and 4 mm. The RPC report provides strong evidence that IMRT implementation is prone to error and that improved quality assurance tools are required. At the heart of radiation therapy dosimetry is the multidimensional dosimeter. However, due to the limited availability of water-equivalent dosimetry materials, research and development in this important field is challenging. In this session, we will review a few dosimeter developments that are either in the laboratory phase or in the pre-commercialization phase. 1) Radiochromic plastic. Novel formulations exhibit light absorbing optical contrast with very little scatter, enabling faster, broad beam optical CT design. 2) Storage phosphor. After irradiation, the dosimetry panels will be read out using a dedicated 2D scanning apparatus in a non-invasive, electro-optic manner and immediately restored for further use. 3) Liquid scintillator. Scintillators convert the energy from x-rays and proton beams into visible light, which can be recorded with a scientific camera (CCD or CMOS) from multiple angles. The 3D shape of the dose distribution can then be reconstructed. 4) Cherenkov emission imaging. Gated intensified imaging allows video-rate passive detection of Cherenkov emission during radiation therapy with the room lights on. Learning Objectives: To understand the physics of a variety of dosimetry techniques based upon optical imaging To investigate the strategies to overcome respective challenges and limitations To explore novel ideas of dosimeter design Supported in part by NIH Grants R01CA148853, R01CA182450, R01CA109558

  14. WE-AB-BRB-03: Real-Time Volumetric Scintillation Dosimetry for Radiation Therapy

    International Nuclear Information System (INIS)

    Beddar, S.

    2016-01-01

    Despite widespread IMRT treatments at modern radiation therapy clinics, precise dosimetric commissioning of an IMRT system remains a challenge. In the most recent report from the Radiological Physics Center (RPC), nearly 20% of institutions failed an end-to-end test with an anthropomorphic head and neck phantom, a test that has rather lenient dose difference and distance-to-agreement criteria of 7% and 4 mm. The RPC report provides strong evidence that IMRT implementation is prone to error and that improved quality assurance tools are required. At the heart of radiation therapy dosimetry is the multidimensional dosimeter. However, due to the limited availability of water-equivalent dosimetry materials, research and development in this important field is challenging. In this session, we will review a few dosimeter developments that are either in the laboratory phase or in the pre-commercialization phase. 1) Radiochromic plastic. Novel formulations exhibit light absorbing optical contrast with very little scatter, enabling faster, broad beam optical CT design. 2) Storage phosphor. After irradiation, the dosimetry panels will be read out using a dedicated 2D scanning apparatus in a non-invasive, electro-optic manner and immediately restored for further use. 3) Liquid scintillator. Scintillators convert the energy from x-rays and proton beams into visible light, which can be recorded with a scientific camera (CCD or CMOS) from multiple angles. The 3D shape of the dose distribution can then be reconstructed. 4) Cherenkov emission imaging. Gated intensified imaging allows video-rate passive detection of Cherenkov emission during radiation therapy with the room lights on. Learning Objectives: To understand the physics of a variety of dosimetry techniques based upon optical imaging To investigate the strategies to overcome respective challenges and limitations To explore novel ideas of dosimeter design Supported in part by NIH Grants R01CA148853, R01CA182450, R01CA109558

  15. Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challenge

    Energy Technology Data Exchange (ETDEWEB)

    Liu, P Z Y; Suchowerska, N; Lambert, J; Abolfathi, P; McKenzie, D R, E-mail: pliu@physics.usyd.edu.au [School of Physics, University of Sydney, NSW 2006 (Australia)

    2011-09-21

    In scintillation dosimetry, a Cerenkov background signal is generated when a conventional fibre optic is exposed to radiation produced by a megavoltage linear accelerator. Three methods of measuring dose in the presence of Cerenkov background are compared. In the first method, a second background fibre is used to estimate the Cerenkov signal in the signal fibre. In the second method, a colour camera is used to measure the combined scintillation and Cerenkov light in two wavelength ranges and a mathematical process is used to extract the scintillation signal. In the third method, a hollow air core light guide is used to carry the scintillation signal through the primary radiation field. In this paper, the strengths and weaknesses of each dosimetry system are identified and recommendations for the optimum method for common clinical dosimetry situations are made.

  16. Scintillation camera with improved output means

    International Nuclear Information System (INIS)

    Lange, K.; Wiesen, E.J.; Woronowicz, E.M.

    1978-01-01

    In a scintillation camera system, the output pulse signals from an array of photomultiplier tubes are coupled to the inputs of individual preamplifiers. The preamplifier output signals are coupled to circuitry for computing the x and y coordinates of the scintillations. A cathode ray oscilloscope is used to form an image corresponding with the pattern in which radiation is emitted by a body. Means for improving the uniformity and resolution of the scintillations are provided. The means comprise biasing means coupled to the outputs of selected preamplifiers so that output signals below a predetermined amplitude are not suppressed and signals falling within increasing ranges of amplitudes are increasingly suppressed. In effect, the biasing means make the preamplifiers non-linear for selected signal levels

  17. BC-454 boron-loaded plastic scintillator

    International Nuclear Information System (INIS)

    Bellian, J.G.

    1984-01-01

    Prototype samples of plastic scintillators containing up to 10% by weight of natural boron have been produced. The maximum size scintillators made to date are 28 mm dia. x 100 mm long. Rods containing up to 2% boron are now made routinely and work is progressing on higher concentrations. The plastics are clear and emit the same blue fluorescence as other common plastic scintillators. It is expected that rods up to 3'' dia. containing 5% boron will be produced during the next few months. BC-454 is particularly useful in neutron research, materials studies, some types of neutron dosimetry, and monitoring of medium to high energy neutrons in the presence of other types radiation. It combines attractive features that enhance its usefulness to the physics community

  18. Optical fiber-applied radiation detection system

    International Nuclear Information System (INIS)

    Nishiura, Ryuichi; Uranaka, Yasuo; Izumi, Nobuyuki

    2001-01-01

    A technique to measure radiation by using plastic scintillation fibers doped radiation fluorescent (scintillator) to plastic optical fiber for a radiation sensor, was developed. The technique contains some superiority such as high flexibility due to using fibers, relatively easy large area due to detecting portion of whole of fibers, and no electromagnetic noise effect due to optical radiation detection and signal transmission. Measurable to wide range of and continuous radiation distribution along optical fiber cable at a testing portion using scintillation fiber and flight time method, the optical fiber-applied radiation sensing system can effectively monitor space radiation dose or apparatus operation condition monitoring. And, a portable type scintillation optical fiber body surface pollution monitor can measure pollution concentration of radioactive materials attached onto body surface by arranging scintillation fiber processed to a plate with small size and flexibility around a man to be tested. Here were described on outline and fundamental properties of various application products using these plastic scintillation fiber. (G.K.)

  19. ANL/WSU radiation damage studies

    International Nuclear Information System (INIS)

    Jankowski, D.; Lopiano, D.; Proudfoot, J.; Underwood, D.; Miles, L.; Neidiger, J.; Tripard, G.

    1993-01-01

    We report preliminary results for the radiation hardness of (polystryrene) plastic scintillator stacks using a spectrum of energy hardened neutrons from a MARK-III TRIGA reactor. The total dose ranged from 100 KRad to 3MRad. The corresponding fluence was 3.8 x 10 13 to 3.8 x 10 14 (n/cm/cm) with the gamma contribution on the order 2--3% (of fluence). The measurements used Li-6, Li-7 Thermo-luminescence dosimeters. Radiochromic/GaF- Chromic film, and activated foils simultaneously allowing an inter-comparison of these various methods of dosimetry

  20. APD Properties and Recovery from Radiation Damage

    CERN Document Server

    Baccaro, Stefania; Caruso, S; Cavallari, Francesca; Dafinei, Ioan; Diemoz, Marcella; Emeliantchik, Igor; Festinesi, Armando; Longo, Egidio; Montecchi, Marco; Organtini, Giovanni; Rosi, G

    1997-01-01

    Avalanche photodiodes will be used to detect scintillation light from PWO crystals in the CMS electromagnetic calorimeter. Properties of Hamamatsu APD are reported special attention has been devoted to the study of radiation hardness and room temperature annealing. We found a fast recovery with a time constant of 1.3 days a medium fast recovery with a lifetime of the order of 10 days and indication of a third component with very long time constant of the order of 300 days.

  1. Applications of a pnCCD detector coupled to columnar structure CsI(Tl) scintillator system in ultra high energy X-ray Laue diffraction

    Science.gov (United States)

    Shokr, M.; Schlosser, D.; Abboud, A.; Algashi, A.; Tosson, A.; Conka, T.; Hartmann, R.; Klaus, M.; Genzel, C.; Strüder, L.; Pietsch, U.

    2017-12-01

    Most charge coupled devices (CCDs) are made of silicon (Si) with typical active layer thicknesses of several microns. In case of a pnCCD detector the sensitive Si thickness is 450 μm. However, for silicon based detectors the quantum efficiency for hard X-rays drops significantly for photon energies above 10 keV . This drawback can be overcome by combining a pixelated silicon-based detector system with a columnar scintillator. Here we report on the characterization of a low noise, fully depleted 128×128 pixels pnCCD detector with 75×75 μm2 pixel size coupled to a 700 μm thick columnar CsI(Tl) scintillator in the photon range between 1 keV to 130 keV . The excellent performance of the detection system in the hard X-ray range is demonstrated in a Laue type X-ray diffraction experiment performed at EDDI beamline of the BESSY II synchrotron taken at a set of several GaAs single crystals irradiated by white synchrotron radiation. With the columnar structure of the scintillator, the position resolution of the whole system reaches a value of less than one pixel. Using the presented detector system and considering the functional relation between indirect and direct photon events Laue diffraction peaks with X-ray energies up to 120 keV were efficiently detected. As one of possible applications of the combined CsI-pnCCD system we demonstrate that the accuracy of X-ray structure factors extracted from Laue diffraction peaks can be significantly improved in hard X-ray range using the combined CsI(Tl)-pnCCD system compared to a bare pnCCD.

  2. Production low cost plastic scintillator by using commercial polystyrene

    International Nuclear Information System (INIS)

    2011-01-01

    Plastic Scintillators can be described as solid materials which contain organic fluorescent compounds dissolved within a polymer matrix. Transparent plastics commonly used for light scintillation are Polystyrene (or PS, poly-vinyl-benzene) and polyvinyl-toluene (or PVT, poly-methyl-styron). By changing the composition of plastic Scintillators some features such as light yield, radiation hardening, decay time etc. can be controlled. Plastic scintillation detectors have been used in nuclear and high energy physics for many decades. Among their benefits are fast response, ease of manufacture and versatility. Their main drawbacks are radiation resistance and cost. Many research projects have concentrated on improving the fundamental properties of plastic scintillators, but little attention has focussed on their cost and easier manufacturing techniques. First plastic Scintillators were produced in 1950's. Activities for production of low cost Scintillators accelerated in second half of 1970's. In 1975 acrylic based Plexipop Scintillator was developed. Despite its low cost, since its structure was not aromatic the light yield of Plexipop was about one quarter of classical Scintillators. Problems arising from slow response time and weak mechanical properties in scintillators developed, has not been solved until 1980. Within the last decade extrusion method became very popular in preparation of low cost and high quality plastic scintillators. In this activity, preliminary studies for low cost plastic scintillator production by using commercial polystyrene pellets and extrusion plus compression method were aimed. For this purpose, PS blocks consist of commercial fluorescent dopant were prepared in June 2008 by use of the extruder and pres in SANAEM. Molds suitable for accoupling to extruder were designed and manufactured and optimum production parameters such as extrusion temperature profile, extrusion rate and moulding pressure were obtained hence, PS Scintillator Blocks

  3. Development of radiation-hard optical links for the CMS tracker at CERN

    International Nuclear Information System (INIS)

    Vasey, F.; Arbet-Engels, V.; Cervelli, G.; Gill, K.; Grabit, R.; Mommaert, C.; Stefanini, G.

    1998-01-01

    A radiation-hard optical link is under development for readout and control of the tracking detector in the future CMS experiment at the CERN Large Hadron Collider. The authors present the optical system architecture based on edge-emitting InGaAsP laser-diode transmitters operating at a wavelength of 1.3 microm, single mode fiber ribbons, multi-way connectors and InGaAsP in photodiode receivers. They report on radiation hardness tests of lasers, photodiodes, fibers and connectors. Increases of laser threshold and pin leakage currents with hadron fluence have been observed together with decreases in laser slope-efficiency and photodiode responsivity. Short lengths of single-mode optical fiber and multi-way connectors have been found to be little affected by radiation damage. They analyze the analog and digital performance of prototype optical links transmitting data generated at a 40 MSample/s rate. Distortion, settling time, bandwidth, noise, dynamic range and bit-error-rate results are discussed

  4. R and D on Radiation Hard Active Media Based on Quartz Plates

    CERN Document Server

    Onel, Yasar

    2014-01-01

    The need for radiation-hard active media in particle detectors is prominently dictated by the future colliders and the future operations of the Large Hadron Collider. The areas of implementation range from calorimetry to beamline instrumentation to specialized forward detectors e.g. luminosity monitors. In this context, we developed the idea of utilizing quartz plates with various surface coating properties as the active medium for such detectors. Plain quartz is a pure Cerenkov radiator which has quite limited photostatistics. In order to improve the efficiency of the photodetection, various methods were investigated including radiation hard wavelength shifters, p-terphenyl or 4pct gallium doped zinc oxide. The readout options include direct coupling of the photodetector to the quartz plate, or fibers. We have studied various geometries and readout options and constructed calorimeter prototypes. Here we report on the results of the previous tests, and the recent developments, which enable several factors of ...

  5. The radiation hardness of silica optical fiber used in the LED-fiber monitor of BLM and BESIII EMC

    International Nuclear Information System (INIS)

    Xue Zhen; Hu Tao; Fang Jian; Xu Zizong; Wang Xiaolian; Lü Junguang; Zhou Li; Cai Xiao; Yu Boxiang; Wang Zhigang; Sun Lijun; Sun Xilei; Zhang Aiwu

    2012-01-01

    LED-fiber system has been used to monitor BLM and BESIII EMC. A radiation hard silica optical fiber is essential for its stability and reliability. Three types of silica optical fibers, silicone-clad silica optical fiber with high OH - content (SeCS), silica-clad silica optical fiber with low OH - content (SCSL) and silica-clad silica opical fiber with high OH - content (SCSH) were studied. In the experiment, 12 groups of fiber samples were irradiated by 60 Co and 3 groups of fiber samples were irradiated by BEPCII background radiation. Radiation hardness: the radiation hardness of SCSH is best and meets the radiation hardness requirement for LED-fiber monitor of BLM and BESIII EMC. The transmission of SeCS and SCSH decreased to around 80% under the 60 Co-irradiation of 5 Gy and 10 Gy, respectively. The radiation hardness of SeCS is worst because of its silicone cladding. Recovery characteristics: 60 Co-irradiated by the same doses, there were both more annealable and more permanent color centers formed in SeCS than SCSL, and for the same kind of fibers, as long as the irradiated doses are under a certain amount (for example, less than 5 Gy for SeCS), the higher the doses, both the more annealable and the more permanent color centers are formed.

  6. Internal Dosimetry Of I-131 For Radiation Workers Based On Analysis Of The Human Urine And Liquid Scintillation Counting

    International Nuclear Information System (INIS)

    Nguyen Van Hung; Pham Hung Thai; Le Van Ngoc

    2011-01-01

    Internal dosimetry of I-131 for radiation workers based on analysis of the human urine, measuring radioactivity by the liquid scintillation system, and dose calculation by the specialized code has been firstly studied at the Nuclear Research Institute. Urine samples from the subjects internally contaminated with I-131 through respiratory ways were collected, chemically processed, measured beta radioactivities of I-131 by the liquid scintillation system of ALOKA-LSC-6100, and then thyroid doses and effective ones for whole-body were calculated by using the specialized code of LUDEP 2.0. Based on chemically separation procedure for I-131 in urine samples and the low background HPGe gamma spectrometer of Canberra for measuring radioactivity, efficiency for chemical separation was determined to be (86.1 ± 5.0)%. The experimental results for 9 subjects with urine samples to be collected during 4 operating courses of Dalat nuclear reactor with production of I-131 (from June to September, 2010) were shown that thyroid doses and effective ones for whole-body for each course of I-131 production were in ranges of from 0.11 to 13.00 mSv and from 0.01 to 0.71 mSv, respectively. Therefore, totally average doses per year for thyroid and whole-body were less than the correlative levels of permissible doses. Besides, the liquid scintillation method was also compared experimentally with the gamma spectrometry (measuring directly urine samples by the gamma spectrometer to be carried out at the Institute before) was shown that errors on dosimetric results between them were less than 12%. This was proved the dosimetry has had a confidence, and it could be applied for internal dosimetry for radiation workers contacting with unsealed sources of I-131 in radiation installations as well as for diagnostic and therapeutic patients in health ones. (author)

  7. Fluorescent organic dyes as radiation converters in scintillation counting and laser techniques

    International Nuclear Information System (INIS)

    Guesten, H.

    1989-01-01

    PMP (1-phenyl-3-mesityl-2-pyrazoline) was selected as color quenching from the category of the sterically hindered 1,3-diphenyl-2-pyrazoline by means of comparative optimization between photo-physical and scintillation-spectroscopic and chemical properties and the costs of synthesis. It is applied in liquid scintillation detectors for the detection of β (T, C-14) and in large-volume liquid scintillators for the detection of neutrinos (Rutherford Lab.). (HP) [de

  8. UV radiation hardness of silicon inversion layer solar cells

    International Nuclear Information System (INIS)

    Hezel, R.

    1990-01-01

    For full utilization of the high spectral response of inversion layer solar cells in the very-short-wavelength range of the solar spectrum sufficient ultraviolet-radiation hardness is required. In addition to the charge-induced passivation achieved by cesium incorporation into the silicon nitride AR coating, in this paper the following means for further drastic reduction of UV light-induced effects in inversion layer solar cells without encapsulation are introduced and interpretations are given: increasing the nitride deposition temperature, silicon surface oxidation at low temperatures, and texture etching and using higher substrate resistivities. High UV radiation tolerance and improvement of the cell efficiency could be obtained simultaneously

  9. HE upgrade beyond phase 1. Finger scintillator option.

    CERN Document Server

    Afanasiev, Sergey; Boyarintsev, A.Yu; Emeliantchik, Igor; Golutvin, Igor; Grinyov, B.V; Ershov, Yuri; Levchuk, Leonid; Litomin, Aliaksandr; Malakhov, Alexander; Moisenz, Petr; Popov, V.F; Shumeiko, Nikolai; Smirnov, Vitaly; Sorokin, Pavlo; Zhmurin, Petro

    2014-01-01

    CMS hadron calorimeters (HB, HE, HO) have been in operation for several years and contributed substantially to the success of the CMS Physics Program. The endcap calorimeter HE suffered more radiation damage than anticipated causing rapid degradation of scintillator segments (tiles) which have a higher radiation flux from secondary particles than HB and HO. A proposal to upgrade of HE calorimeter will provide a solution for survivability at future LHC higher luminosity. A finger-strip plastic scintillator option has many advantages and is a lower cost alternative to keep the excellent HE performance at high luminosity. Measurements and simulations have been performed and this method is a good upgrade strategy.

  10. Improvement of optical properties and radiation hardness of NaBi(WO sub 4) sub 2 Cherenkov crystals

    CERN Document Server

    Zadneprovski, B I; Polyansky, E V; Devitsin, E G; Kozlov, V A; Potashov, S Yu; Terkulov, A R

    2002-01-01

    On the basis of the data on melt evaporation while growing NaBi(WO sub 4) sub 2 Cherenkov crystals, the formation of nonstoichiometry and most probable types of dot defects of the crystals have been considered. The influence of melt nonstoichiometry and doping with Sc on optical transmission and radiation hardness of the crystals has been experimentally investigated. The surplus of WO sub 3 has been established to increase optical transmission and radiation hardness and lack of Bi sub 2 O sub 3 in the melt to reduce radiation hardness. Sc doping is shifting the absorption edge to UV region by 30-35 nm and is increasing radiation hardness of the crystals about three-fold. Analytical estimations give the increase of the number of Cherenkov photons by a factor of 1.3, which leads to an improvement of the energy resolution of a calorimeter based on NaBi(WO sub 4) sub 2 :Sc crystals compared with undoped NaBi(WO sub 4) sub 2 of approximately 15%.

  11. Detection of high energy gamma radiations with liquid rare gases as scintillators; Detection des rayonnements Gamma de grande energie avec les gaz rares liquides comme scintillateurs

    Energy Technology Data Exchange (ETDEWEB)

    Ho, Phan Xuan

    1965-11-25

    This research thesis reports the study of a sensor based on a liquid scintillator for the detection of high energy (10 to 30 MeV) gamma radiations. The scintillator is a liquefied argon or xenon rare gas. The author first studies the process of energy transfer from the particle to the sensing medium. He addresses the different involved elements and phenomena: electromagnetic radiations (Compton Effect, photoelectric effect, pair production, and total gamma absorption), charged particles (braking radiation, collisions) and application to gamma spectrometry. He describes and discusses the scintillation mechanisms (scintillation of organic and inorganic materials), the general characteristics of scintillators (impurities, converters), and then reports the practical realisation of the sensor. Results are presented and discussed [French] Dans ce travail, nous nous proposons d'etudier une technique. Il s'agit d'un detecteur a scintillateur liquide pour la detection des rayonnements gamma energiques (10 a 30 MeV). Le scintillateur utilise est un gaz rare liquefie argon ou xenon. Nous examinerons d'abord les processus de transfert de l'energie de la particule au milieu detecteur puis les mecanismes de scintillation en general pour pouvoir exploiter au mieux les phenomenes favorables. Nous presenterons ensuite la realisation pratique du detecteur. Ses qualites (et defauts) trouveront leur place dans la fin de ce memoire. Bien qu'a l'heure actuelle, par la methode de Kyropoulos, on puisse faire pousser des gros cristaux d'iodure de sodium, l'utilisation des 'gaz rares' liquefies comme scintillateurs est, grace a la brievete de la scintillation, tres utile lorsqu'on recherche un fort taux de comptage (jusqu'a 10 impulsions par seconde) ou lorsqu'on veut resoudre certains problemes de coincidence. Les cristaux NaI(Tl) de grandes dimensions sont d'un montage facile mais leur manipulation requiert beaucoup de precautions du fait qu'ils supportent tres mal les chocs thermiques

  12. Scintillation densimeter for liquids and an isotopic conveyor weighers with plastic scintillator

    International Nuclear Information System (INIS)

    Makhaj, B.; Antonyak, V.; Plyater, Z.

    1979-01-01

    The method is described of the weighted material's mass measuring according to the results of the conveyor momentary load measurement derived from the attenuation of radiation in the transmission geometry, conveyor belt velocity measurement and digital processing of the signals from the measurement of the bouth values. In the measuring gage there are located: the point type gamma source of cesium-137 with 4 mCi capacity, the scintillation detector with plastic cylindric scintillator of 5 cm in diameter and with the length approximately equal to the width of the conveyor belt and also the tachometer-generator. The conveyor weighers described is intended for use with conveyor having belt; from 60 to 180 cm wide. The results are given of industrial exploitation of the instrument [ru

  13. Spectral Analysis Method of Plastic Scintillator-based Radiation Detector against Nuclear/Radiological Terrorism

    International Nuclear Information System (INIS)

    Kwak, Sung-Woo; Yoo, Ho-Sik; Jang, Sung Soon; Kim, Jung-Soo; Yoon, Wan-Ki

    2008-01-01

    In these days, the threats relating to nuclear or radioactive materials have become a matter of internationally increased grave concern. A plastic scintillation detector in radiation portal monitoring (RPM) application has been used to detect radioactive sources in steel scrap entering reprocessing facilities, and to detect illicit transport of radioactive material across border ports-of-entry. The detection systems for RPM application usually are large and can not easily be moved to a different location. For some situations, an inconspicuous and mobile system for the radioactive or nuclear material during road transport is needed. The mobile radiation detection system has employed a NaI- based radiation detector to detect and identify the material hidden in vehicle. There are some operational constraints - short measuring time, weak activity due to heavy shield of illegal source, long distance - of inspection system in such nuclear security applications. Due to these constraints, large area sensor is required to maximize its sensitivity. Large NaI material, however, is extremely expensive. In designing a radiation detector for prevention of illicit trafficking of nuclear or radioactive materials, the trade-off should be carefully optimized between performance and cost in order to achieve cost-effective inspection system. For the cost-effective mobile radiation detection system, this paper describes new spectral analysis method to use the crude spectroscopic information available from a plastic detector to discriminate other man-made radiation source from NORM

  14. Radiation hardness of CMS pixel barrel modules

    CERN Document Server

    Rohe, T; Erdmann, W; Kästli, H C; Khalatyan, S; Meier, B; Radicci, V; Sibille, J

    2010-01-01

    Pixel detectors are used in the innermost part of the multi purpose experiments at LHC and are therefore exposed to the highest fluences of ionising radiation, which in this part of the detectors consists mainly of charged pions. The radiation hardness of all detector components has thoroughly been tested up to the fluences expected at the LHC. In case of an LHC upgrade, the fluence will be much higher and it is not yet clear how long the present pixel modules will stay operative in such a harsh environment. The aim of this study was to establish such a limit as a benchmark for other possible detector concepts considered for the upgrade. As the sensors and the readout chip are the parts most sensitive to radiation damage, samples consisting of a small pixel sensor bump-bonded to a CMS-readout chip (PSI46V2.1) have been irradiated with positive 200 MeV pions at PSI up to 6E14 Neq and with 21 GeV protons at CERN up to 5E15 Neq. After irradiation the response of the system to beta particles from a Sr-90 source w...

  15. Scintillation camera for high activity sources

    International Nuclear Information System (INIS)

    Arseneau, R.E.

    1978-01-01

    The invention described relates to a scintillation camera used for clinical medical diagnosis. Advanced recognition of many unacceptable pulses allows the scintillation camera to discard such pulses at an early stage in processing. This frees the camera to process a greater number of pulses of interest within a given period of time. Temporary buffer storage allows the camera to accommodate pulses received at a rate in excess of its maximum rated capability due to statistical fluctuations in the level of radioactivity of the radiation source measured. (U.K.)

  16. Radiation hardness properties of full-3D active edge silicon sensors

    Czech Academy of Sciences Publication Activity Database

    Da Via, C.; Hasi, J.; Kenney, C.; Linhart, V.; Parker, S.; Slavíček, T.; Watts, S. J.; Bém, Pavel; Horažďovský, T.; Pospíšil, S.

    2008-01-01

    Roč. 587, 2-3 (2008), s. 243-249 ISSN 0168-9002 Institutional research plan: CEZ:AV0Z10480505 Keywords : silicon detectors * radiation hardness * 3D Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders Impact factor: 1.019, year: 2008

  17. Examination of the Properties of a Spent Fuel based Electricity Generation System - Scintillator Performance Analysis

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Haneol; Yim, Man-Sung [KAIST, Daejeon (Korea, Republic of)

    2016-05-15

    Gammavoltaic was proposed by Karl Scharf in 1960. The low efficiency resulted in gammavoltaic being used as a radiation detector. In the 1990s the efficiency of gammavoltaic increased by the use of a scintillator. Gammavoltaic was further studied as a power source for spent fuel transportation and a nuclear battery in the 2000s Haneol Lee and Man-Sung Yim also suggested electricity generation system based on spent fuel stored inside the fuel pool of a nuclear power plant. This study proposed the systematic design of an electricity conversion system using CsI(Tl) scintillator and a-Si photovoltaic cell. As such, this study is selected to be a reference paper. The results of this paper indicate a self-absorption effect from the reference model. This effect is negligible while the irradiation degradation has to be considered. Two main ways to reduce radiation induced degradation are scintillator shielding and replacing scintillator material with a material having higher radiation resistance. The analysis of the scintillator used in the 'electricity generation system using gamma radiation from spent fuel' was performed to evaluate the ideal electricity generation in the reference research.

  18. Examination of the Properties of a Spent Fuel based Electricity Generation System - Scintillator Performance Analysis

    International Nuclear Information System (INIS)

    Lee, Haneol; Yim, Man-Sung

    2016-01-01

    Gammavoltaic was proposed by Karl Scharf in 1960. The low efficiency resulted in gammavoltaic being used as a radiation detector. In the 1990s the efficiency of gammavoltaic increased by the use of a scintillator. Gammavoltaic was further studied as a power source for spent fuel transportation and a nuclear battery in the 2000s Haneol Lee and Man-Sung Yim also suggested electricity generation system based on spent fuel stored inside the fuel pool of a nuclear power plant. This study proposed the systematic design of an electricity conversion system using CsI(Tl) scintillator and a-Si photovoltaic cell. As such, this study is selected to be a reference paper. The results of this paper indicate a self-absorption effect from the reference model. This effect is negligible while the irradiation degradation has to be considered. Two main ways to reduce radiation induced degradation are scintillator shielding and replacing scintillator material with a material having higher radiation resistance. The analysis of the scintillator used in the 'electricity generation system using gamma radiation from spent fuel' was performed to evaluate the ideal electricity generation in the reference research

  19. RD50 recent results: Development of radiation hard sensors for SLHC

    CERN Document Server

    Macchiolo, Anna

    2009-01-01

    The need for radiation hard semiconductor detectors for the tracker regions in high energy physics experiments at a future high luminosity hadron collider, like the proposed LHC upgrade, has led to the formation of the CERN RD50 collaboration. The R&D directions of RD50 follow two paths: the optimization of radiation hard bulk materials (Material Engineering) and the development of new detector designs (Device Engineering) as 3D sensors, thin sensors and n-in-p sensors. Some of the RD50 most recent results about silicon detectors are reported in this paper, with special reference to: (i) identification of defects responsible for long term annealing, (ii) charge collection efficiency of irradiated planar devices, in particular n-in-p microstrip detectors and epitaxial diodes, (iii) charge collection efficiency of double-type column 3D detectors, (iv) comparison of the performances of FZ and MCZ structures under mixed irradiation.

  20. Scintillation detectors, applications in ambient dose rate measurement of ionizing radiation

    International Nuclear Information System (INIS)

    Harja, C.

    1997-01-01

    For the task of developing an ambient dose rate measuring probe, the company MAB decided for a design combining an organic plastic scintillator and a secondary electron multiplier. MAB tested a range of available plastic scintillators suitable for this task, and one proved to be particularly good for the intended purpose. This scintillator is a product of the British company NE Technology Ltd. and was specially developed for dosimetry applications, offering the following advantages: The response within the range from 33 keV to 1.3 MeV (Co-60) is much more constant than with combined-design types; there is no afterglow under conditions of excessive load. (orig./CB) [de

  1. SENTIRAD-An innovative personal radiation detector based on a scintillation detector and a silicon photomultiplier

    International Nuclear Information System (INIS)

    Osovizky, A.; Ginzburg, D.; Manor, A.; Seif, R.; Ghelman, M.; Cohen-Zada, I.; Ellenbogen, M.; Bronfenmakher, V.; Pushkarsky, V.; Gonen, E.; Mazor, T.; Cohen, Y.

    2011-01-01

    The alarming personal radiation detector (PRD) is a device intended for Homeland Security (HLS) applications. This portable device is designed to be worn or carried by security personnel to detect photon-emitting radioactive materials for the purpose of crime prevention. PRD is required to meet the scope of specifications defined by various HLS standards for radiation detection. It is mandatory that the device be sensitive and simultaneously small, pocket-sized, of robust mechanical design and carriable on the user's body. To serve these specialized purposes and requirements, we developed the SENTIRAD, a new radiation detector designed to meet the performance criteria established for counterterrorist applications. SENTIRAD is the first commercially available PRD based on a CsI(Tl) scintillation crystal that is optically coupled with a silicon photomultiplier (SiPM) serving as a light sensor. The rapidly developing technology of SiPM, a multipixel semiconductor photodiode that operates in Geiger mode, has been thoroughly investigated in previous studies. This paper presents the design considerations, constraints and radiological performance relating to the SENTIRAD radiation sensor.

  2. Radiation hardness of Ce-doped sol-gel silica fibers for high energy physics applications.

    Science.gov (United States)

    Cova, Francesca; Moretti, Federico; Fasoli, Mauro; Chiodini, Norberto; Pauwels, Kristof; Auffray, Etiennette; Lucchini, Marco Toliman; Baccaro, Stefania; Cemmi, Alessia; Bártová, Hana; Vedda, Anna

    2018-02-15

    The results of irradiation tests on Ce-doped sol-gel silica using x- and γ-rays up to 10 kGy are reported in order to investigate the radiation hardness of this material for high-energy physics applications. Sol-gel silica fibers with Ce concentrations of 0.0125 and 0.05 mol. % are characterized by means of optical absorption and attenuation length measurements before and after irradiation. The two different techniques give comparable results, evidencing the formation of a main broad radiation-induced absorption band, peaking at about 2.2 eV, related to radiation-induced color centers. The results are compared with those obtained on bulk silica. This study reveals that an improvement of the radiation hardness of Ce-doped silica fibers can be achieved by reducing Ce content inside the fiber core, paving the way for further material development.

  3. Water-equivalent plastic scintillation detectors for radiotherapy dosimetry

    International Nuclear Information System (INIS)

    Beddar, A.S.

    1995-01-01

    Plastic scintillation dosimetry is a promising new method of measuring absorbed dose for high energy radiotherapy beams. The theory behind this concept will be presented along with the many advantages that it offers over conventional dosimetry. A variety of plastic scintillation detector systems have been recently developed for photon and electron dosimetry. These new water-equivalent detectors use small to miniature plastic scintillators. Their attractive feature lies in their use for field mapping in water, particularly for small fields, high dose gradient regions, and near inhomogeneous interfaces, or for in-vivo insertions. The physical characteristics and the dosimetric properties of these scintillators will be presented, discussed, and compared to the commonly used detectors in radiation dosimetry. The system first used successfully for multi-purpose radiotherapy field mapping, as well as other systems, will be described. The technical challenges of the design of these detectors including the optical coupling to small fibers will be discussed. One of the limitations, at the present time, is the radiation-induced light produced in the optical fibers that are used to transmit the signal to the photodetectors. The mechanisms of these spurious effects will be identified and discussed with emphasis on signal-to-noise improvements

  4. An analytical model of nonproportional scintillator light yield in terms of recombination rates

    International Nuclear Information System (INIS)

    Bizarri, G.; Moses, W. W.; Singh, J.; Vasil'ev, A. N.; Williams, R. T.

    2009-01-01

    Analytical expressions for the local light yield as a function of the local deposited energy (-dE/dx) and total scintillation yield integrated over the track of an electron of initial energy E are derived from radiative and/or nonradiative rates of first through third order in density of electronic excitations. The model is formulated in terms of rate constants, some of which can be determined independently from time-resolved spectroscopy and others estimated from measured light yield efficiency as a constraint assumed to apply in each kinetic order. The rates and parameters are used in the theory to calculate scintillation yield versus primary electron energy for comparison to published experimental results on four scintillators. Influence of the track radius on the yield is also discussed. Results are found to be qualitatively consistent with the observed scintillation light yield. The theory can be applied to any scintillator if the rates of the radiative and nonradiative processes are known

  5. A Radiation-Hard Analog Memory In The AVLSI-RA Process

    International Nuclear Information System (INIS)

    Britton, C.L. Jr.; Wintenberg, A.L.; Read, K.F.; Simpson, M.L.; Young, G.R.; Clonts, L.G.; Kennedy, E.J.; Smith, R.S.; Swann, B.K.; Musser, J.A.

    1995-01-01

    A radiation hardened analog memory for an Interpolating Pad Camber has been designed at Oak Ridge National Laboratory and fabricated by Harris Semiconductor in the AVLSI-RA CMOS process. The goal was to develop a rad-hard analog pipeline that would deliver approximately 9-bit performance, a readout settling time of 500ns following read enable, an input and output dynamic range of +/-2.25V, a corrected rms pedestal of approximately 5mV or less, and a power dissipation of less than 10mW/channel. The pre- and post-radiation measurements to 5MRad are presented

  6. Radiation damages in chemical components of organic scintillator detectors

    International Nuclear Information System (INIS)

    Fernandes Neto, Jose Maria

    2003-01-01

    Samples containing PPO (1%, g/ml), diluted in toluene, they were irradiated in a 60 Co irradiator (6.46 kGy/h) at different doses. The PPO concentration decay bi-exponentially with the dose, generating the degradation products: benzoic acid, benzamide and benzilic alcohol. The liquid scintillator system was not sensitive to the radiation damage until 20 kGy. Otherwise, the pulse height analysis showed that dose among 30 to 40 kGy generate significant loss of quality of the sensor (liquid scintillating) and the light yield was reduced in half with the dose of (34.04 ± 0.80) kGy. This value practically was confirmed by the photo peak position analysis that resulted D 1/2 = (31.7 ± 1,4) kGy, The transmittance, at 360 nm, of the irradiated solution decreased exponentially. The compartmental model using five compartments (fast decay PPO, slow decay PPO, benzamide, benzoic acid and benzilic alcohol) it was satisfactory to explain the decay of the PPO in its degradation products in function of the dose. The explanation coefficient r 2 = 0.985636 assures that the model was capable to explain 98.6% of the experimental variations. The Target Theory together with the Compartmental Analysis showed that PPO irradiated in toluene solution presents two sensitive molecular diameters both of them larger than the true PPO diameter. >From this analysis it showed that the radiolytic are generated, comparatively, at four toluene molecules diameter far from PPO molecules. For each one PPO-target it was calculated the G parameter (damage/100 eV). For the target expressed by the fast decay the G value was (418.4 ± 54.1) damages/100 eV, and for the slow decay target the G value was (54.5 ± 8.9) damages/100 eV. The energies involved in the chemical reactions were w (0.239 ± 0.031) eV/damage (fast decay) and w = (1 834 ± 0.301) eV/damage (slow decay). (author)

  7. Radiation hard silicon particle detectors for HL-LHC—RD50 status report

    Energy Technology Data Exchange (ETDEWEB)

    Terzo, S., E-mail: Stefano.Terzo@mpp.mpg.de

    2017-02-11

    It is foreseen to significantly increase the luminosity of the LHC by upgrading towards the HL-LHC (High Luminosity LHC). The Phase-II-Upgrade scheduled for 2024 will mean unprecedented radiation levels, way beyond the limits of the silicon trackers currently employed. All-silicon central trackers are being studied in ATLAS, CMS and LHCb, with extremely radiation hard silicon sensors to be employed on the innermost layers. Within the RD50 Collaboration, a massive R&D program is underway across experimental boundaries to develop silicon sensors with sufficient radiation tolerance. We will present results of several detector technologies and silicon materials at radiation levels corresponding to HL-LHC fluences. Based on these results, we will give recommendations for the silicon detectors to be used at the different radii of tracking systems in the LHC detector upgrades. In order to complement the measurements, we also perform detailed simulation studies of the sensors. - Highlights: • The RD50 collaboration investigates the radiation hardness of silicon sensors. • Different approaches to simulate the detector response after irradiation are shown. • HV-CMOS are cost-effective solution for the outer pixel layers at HL-LHC. • 3D and thin planar sensors with slim edges are solutions for innermost layers at HL-LHC. • Sensors with intrinsic gain are investigated to develop ultra-fast silicon detectors.

  8. Collimated trans-axial tomographic scintillation camera

    International Nuclear Information System (INIS)

    1980-01-01

    The objects of this invention are first to reduce the time required to obtain statistically significant data in trans-axial tomographic radioisotope scanning using a scintillation camera. Secondly, to provide a scintillation camera system to increase the rate of acceptance of radioactive events to contribute to the positional information obtainable from a known radiation source without sacrificing spatial resolution. Thirdly to reduce the scanning time without loss of image clarity. The system described comprises a scintillation camera detector, means for moving this in orbit about a cranial-caudal axis relative to a patient and a collimator having septa defining apertures such that gamma rays perpendicular to the axis are admitted with high spatial resolution, parallel to the axis with low resolution. The septa may be made of strips of lead. Detailed descriptions are given. (U.K.)

  9. Scintillators for fiber optics: system sensitivity and bandwidth as a function of fiber length

    International Nuclear Information System (INIS)

    Lutz, S.S.; Franks, L.A.; Fluornoy, J.M.; Lyons, P.B.

    1981-01-01

    Scintillators have been employed for several years as ionizing radiation-to-light converters in plasma diagnostic experiments that utilize fiber optics. Until recently, nanosecond and subnanosecond scintillators were available only in the near ultraviolet. However, the bandwidth and transmission properties of fiber optics both strongly favor operation at longer wavelengths. More recently, nanosecond and subnanosecond scintillators with emission peaks around 480 nm have been reported. A time-resolved plasma-imaging experiment using one of these scintillators and 100 channels of graded-index fiber, each 500 m long, has been successfully tested on a nuclear event at the Nevada Test Site. During the past year we have developed several new scintillator systems with emission wavelengths more compatible with fiber optics and with response times in the nanosecond and subnanosecond time region. One scintillator, based on Kodak dye 14567 (DCM), has an emission maximum at 650 nm and a response time (FWHM) of 1.2 ns. Experimental data on system sensitivity and bandwidth versus fiber length are presented for three fluor-fiber systems. Data on fluor formulation, response time, and linearity-of-response are given, and a model for scintillator nonlinearity, based on solvent, radiation-induced, transient absorption, is presented

  10. Evaluation of dual γ-ray imager with active collimator using various types of scintillators.

    Science.gov (United States)

    Lee, Wonho; Lee, Taewoong; Jeong, Manhee; Kim, Ho Kyung

    2011-10-01

    The performance of a specialized dual γ-ray imager using both mechanical and electronic collimation was evaluated by Monte Carlo simulation (MCNP5). The dual imager consisted of an active collimator and a planar detector that were made from scintillators. The active collimator served not only as a coded aperture for mechanical collimation but also as a first detector for electronic collimation. Therefore, a single system contained both mechanical and electronic collimation. Various types of scintillators were tested and compared with each other in terms of their angular resolution, efficiency, and background noise. In general, a BGO active collimator had the best mechanical collimation performance, and an LaCl₃(Ce) active collimator provided the best electronic collimation performance. However, for low radiation energies, the mechanical collimation images made from both scintillators showed the same quality, and, for high radiation energies, electronic collimation images made from both scintillators also show similar quality. Therefore, if mechanical collimation is used to detect low-energy radiation and electronic collimation is applied to reconstruct a high-energy source, either LaCl₃(Ce) or BGO would be appropriate for the active collimator of a dual γ-ray imager. These results broaden the choice of scintillators for the active collimator of the dual γ-ray imager, which makes it possible to consider other factors, such as machinability and cost, in making the imager. As a planar detector, BGO showed better performance than other scintillators since its radiation detection efficiency was highest of all. Copyright © 2011 Elsevier Ltd. All rights reserved.

  11. Study of runaway electrons using dosimetry of hard x-ray radiations in Damavand tokamak

    Energy Technology Data Exchange (ETDEWEB)

    Rasouli, C.; Pourshahab, B.; Rasouli, H. [Plasma Physics and Nuclear Fusion Research School, Nuclear Science and Technology Research Institute, AEOI, PO Box 14155-1339, Tehran (Iran, Islamic Republic of); Hosseini Pooya, S. M.; Orouji, T. [Radiation Application Research School, Nuclear Science and Technology Research Institute, AEOI, PO Box 14155-1339, Tehran (Iran, Islamic Republic of)

    2014-05-15

    In this work several studies have been conducted on hard x-ray emissions of Damavand tokamak based on radiation dosimetry using the Thermoluminescence method. The goal was to understand interactions of runaway electrons with plasma particles, vessel wall, and plasma facing components. Total of 354 GR-200 (LiF:Mg,Cu,P) thermoluminescence dosimeter (TLD) crystals have been placed on 118 points – three TLDs per point – to map hard x-ray radiation doses on the exterior of the vacuum vessel. Results show two distinctive levels of x-ray radiations doses on the exterior of the vessel. The low-dose area on which measured dose is about 0.5 mSv/shot. In the low-dose area there is no particular component inside the vessel. On the contrary, on high-dose area of the vessel, x-ray radiations dose exceeds 30 mSv/shot. The high-dose area coincides with the position of limiters, magnetic probe ducts, and vacuum vessel intersections. Among the high-dose areas, the highest level of dose is measured in the position of the limiter, which could be due to its direct contact with the plasma column and with runaway electrons. Direct collisions of runaway electrons with the vessel wall and plasma facing components make a major contribution for production of hard x-ray photons in Damavand tokamak.

  12. A Demonstrator Analog Signal Processing Circuit in a Radiation Hard SOI-CMOS Technology

    CERN Multimedia

    2002-01-01

    % RD-9 A Demonstrator Analog Signal Processing Circuit in a Radiation Hard SOI-CMOS Technology \\\\ \\\\Radiation hardened SOI-CMOS (Silicon-On-Insulator, Complementary Metal-Oxide- \\linebreak Semiconductor planar microelectronic circuit technology) was a likely candidate technology for mixed analog-digital signal processing electronics in experiments at the future high luminosity hadron colliders. We have studied the analog characteristics of circuit designs realized in the Thomson TCS radiation hard technologies HSOI3-HD. The feature size of this technology was 1.2 $\\mu$m. We have irradiated several devices up to 25~Mrad and 3.10$^{14}$ neutrons cm$^{-2}$. Gain, noise characteristics and speed have been measured. Irradiation introduces a degradation which in the interesting bandwidth of 0.01~MHz~-~1~MHz is less than 40\\%. \\\\ \\\\Some specific SOI phenomena have been studied in detail, like the influence on the noise spectrum of series resistence in the thin silicon film that constitutes the body of the transistor...

  13. Development of scintillation materials for medical imaging and other applications

    International Nuclear Information System (INIS)

    Melcher, C. L.

    2013-01-01

    Scintillation materials that produce pulses of visible light in response to the absorption of energetic photons, neutrons, and charged particles, are widely used in various applications that require the detection of radiation. The discovery and development of new scintillators has accelerated in recent years, due in large part to their importance in medical imaging as well as in security and high energy physics applications. Better understanding of fundamental scintillation mechanisms as well as the roles played by defects and impurities have aided the development of new high performance scintillators for both gamma-ray and neutron detection. Although single crystals continue to dominate gamma-ray based imaging techniques, composite materials and transparent optical ceramics potentially offer advantages in terms of both synthesis processes and scintillation performance. A number of promising scintillator candidates have been identified during the last few years, and several are currently being actively developed for commercial production. Purification and control of raw materials and cost effective crystal growth processes can present significant challenges to the development of practical new scintillation materials.

  14. Study of the correlation of scintillation decay and emission wavelength

    International Nuclear Information System (INIS)

    Yanagida, Takayuki; Fujimoto, Yutaka; Yamaji, Akihiro; Kawaguchi, Noriaki; Kamada, Kei; Totsuka, Daisuke; Fukuda, Kentaro; Yamanoi, Kohei; Nishi, Ryosuke; Kurosawa, Shunsuke; Shimizu, Toshihiko; Sarukura, Nobuhiko

    2013-01-01

    In photoluminescence which directly excites the emission center of phosphor material is known to have a correlation between the emission wavelength and the decay time based on quantum mechanics. In scintillation phenomenon, host lattice of the material is first excited by ionizing radiation and then the excitation energy is transferred to emission centers. For the first time, we investigated the correlation between the scintillation decay and the emission wavelength by using pulse X-ray equipped streak camera system which could observe time and wavelength resolved scintillation phenomenon. Investigated materials were Ce 3+ , Pr 3+ and Nd 3+ doped oxides and fluorides which all showed 5d-4f transition based emission. As a result, we obtained the relation that τ (scintillation decay time) was proportional to the λ 2.15 (emission wavelength). -- Highlights: ► The correlation between emission wavelength and scintillation decay time is investigated. ► Photoluminescence decay times are also evaluated and compared with scintillation decay times. ► It is proved the relaxation process in emission center is dominant even in scintillation decay

  15. Excitonic processes and their contribution to non-proportionality observed in the light yield of inorganic scintillators

    International Nuclear Information System (INIS)

    Singh, J.; Koblov, A.

    2013-01-01

    Using the derived expression for the light yield in a scintillator, the influence of linear radiative and non-radiative (quenching) rates on the non-proportionality in light yield is studied. It is found that if the excitation created within the electron track initiated by a γ-photon incident on a scintillator, remains mainly excitonic, then non-proportionality can be minimized by inventing a scintillator material with linear radiative rate > 10 7 s -1 , linear quenching rate 6 s -1 and track radius ≥ 70 nm along with maintaining the rates of other non-linear processes as discovered earlier. If one can increase the linear radiative rate to 10 9 s -1 , then the non-proportionality can be eliminated at a track radius > 20 nm. (authors)

  16. Scintillation counter, segmented shield

    International Nuclear Information System (INIS)

    Olson, R.E.; Thumim, A.D.

    1975-01-01

    A scintillation counter, particularly for counting gamma ray photons, includes a massive lead radiation shield surrounding a sample-receiving zone. The shield is disassembleable into a plurality of segments to allow facile installation and removal of a photomultiplier tube assembly, the segments being so constructed as to prevent straight-line access of external radiation through the shield into radiation-responsive areas. Provisions are made for accurately aligning the photomultiplier tube with respect to one or more sample-transmitting bores extending through the shield to the sample receiving zone. A sample elevator, used in transporting samples into the zone, is designed to provide a maximum gamma-receiving aspect to maximize the gamma detecting efficiency. (U.S.)

  17. Dosimetry in clinical static magnetic fields using plastic scintillation detectors

    DEFF Research Database (Denmark)

    Stefanowicz, S.; Latzel, H.; Lindvold, Lars René

    2013-01-01

    . In conclusion, we found some deviations up to 7% of the supposed signal. Although the scintillators are of much denser material, we measured the same behavior in signal as in (Meijsing et al., 2009) for a Farmer ionization chamber or as in (Raaijmakers et al., 2007) for films described which indicates radiation......-vivo dosimetry in radiation treatments and diagnostics and could be, being all-optical, promising candidates for this application. To study the basic feasibility of using PSDs with organic scintillators in magnetic fields, we measured the response of these dosimeters in presence of magnetic fields up to 1 T...

  18. Study of the response of a lithium yttrium borate scintillator based neutron rem counter by Monte Carlo radiation transport simulations

    Science.gov (United States)

    Sunil, C.; Tyagi, Mohit; Biju, K.; Shanbhag, A. A.; Bandyopadhyay, T.

    2015-12-01

    The scarcity and the high cost of 3He has spurred the use of various detectors for neutron monitoring. A new lithium yttrium borate scintillator developed in BARC has been studied for its use in a neutron rem counter. The scintillator is made of natural lithium and boron, and the yield of reaction products that will generate a signal in a real time detector has been studied by FLUKA Monte Carlo radiation transport code. A 2 cm lead introduced to enhance the gamma rejection shows no appreciable change in the shape of the fluence response or in the yield of reaction products. The fluence response when normalized at the average energy of an Am-Be neutron source shows promise of being used as rem counter.

  19. Study of the response of a lithium yttrium borate scintillator based neutron rem counter by Monte Carlo radiation transport simulations

    Energy Technology Data Exchange (ETDEWEB)

    Sunil, C., E-mail: csunil11@gmail.com [Accelerator Radiation Safety Section, Health Physics Division, Bhabha Atomic Research Centre, Mumbai 400085 (India); Tyagi, Mohit [Technical Physics Division, Bhabha Atomic Research Centre, Mumbai 400085 (India); Biju, K.; Shanbhag, A.A.; Bandyopadhyay, T. [Accelerator Radiation Safety Section, Health Physics Division, Bhabha Atomic Research Centre, Mumbai 400085 (India)

    2015-12-11

    The scarcity and the high cost of {sup 3}He has spurred the use of various detectors for neutron monitoring. A new lithium yttrium borate scintillator developed in BARC has been studied for its use in a neutron rem counter. The scintillator is made of natural lithium and boron, and the yield of reaction products that will generate a signal in a real time detector has been studied by FLUKA Monte Carlo radiation transport code. A 2 cm lead introduced to enhance the gamma rejection shows no appreciable change in the shape of the fluence response or in the yield of reaction products. The fluence response when normalized at the average energy of an Am–Be neutron source shows promise of being used as rem counter.

  20. Low cost plastic scintillator by using commercial polystyrene

    International Nuclear Information System (INIS)

    Oktar, O.; Ari, G.; Guenduez, O.; Demirel, H.; Demirbas, A.

    2009-01-01

    Plastic scintillation detectors have been used in nuclear and high energy physics for many decades. Among their benefits are fast response, ease of manufacture and versatility. Their main drawbacks are radiation resistance and cost. Plastic Scintillators can be described as solid materials which contain organic fluorescent compounds dissolved within a polymer matri10. Transparent plastics commonly used for light scintillation are Polystyrene (or PS, poly-vinyl-benzene) and polyvinyl-toluene (or PVT, poly-methyl-styron). In this activity, preliminary studies for low cost plastic scintillator production by using commercial polystyrene pellets and extrusion method were aimed. For this purpose, PS blocks consist of commercial fluorescent dopant were prepared by an extruder in SANAEM. Molds suitable for extruder were designed and manufactured and optimum production parameters such as extrusion temperature profile, extrusion rate and pressure were obtained. Plastic blocks prepared were optically and mechanically tested and its response against various radioactive sources was measured.This study has shown that plastic scintillators imported can be produced in SANAEM domestically and be used for detection of radioactive materials within the country or border gates.

  1. Life evaluation of FR-CV cable on thermal-radiation combined aging by micro-hardness

    International Nuclear Information System (INIS)

    Sugiyama, Masahiko; Ogata, Akimasa; Nitta, Makoto; Tani, Tsuneo; Yagi, Toshiaki; Seguchi, Tadao.

    1996-01-01

    For the evaluation of cable life for the application to nuclear facilities, the accelerated test was conducted by the combination of radiation and thermal oxidation. The degradation of FR-CV cable by the aging was monitored by tensile test, micro-hardness test, and gel-fraction measurement. The micro-hardness increased with the progress of degradation and related well with decrease of ultimate elongation of the sheath material, and was also reflected by the loss of plasticizer. The micro-hardness technique has a possibility to detect the degradation of cable as a non-destructive detector. (author)

  2. Final LDRD report : advanced plastic scintillators for neutron detection.

    Energy Technology Data Exchange (ETDEWEB)

    Vance, Andrew L.; Mascarenhas, Nicholas; O' Bryan, Greg; Mrowka, Stanley

    2010-09-01

    This report summarizes the results of a one-year, feasibility-scale LDRD project that was conducted with the goal of developing new plastic scintillators capable of pulse shape discrimination (PSD) for neutron detection. Copolymers composed of matrix materials such as poly(methyl methacrylate) (PMMA) and blocks containing trans-stilbene (tSB) as the scintillator component were prepared and tested for gamma/neutron response. Block copolymer synthesis utilizing tSBMA proved unsuccessful so random copolymers containing up to 30% tSB were prepared. These copolymers were found to function as scintillators upon exposure to gamma radiation; however, they did not exhibit PSD when exposed to a neutron source. This project, while falling short of its ultimate goal, demonstrated the possible utility of single-component, undoped plastics as scintillators for applications that do not require PSD.

  3. Single-photoelectron noise reduction in scintillation detectors

    International Nuclear Information System (INIS)

    Marvin, T.P.

    1995-10-01

    The 1994--95 search at SLAC for mulicharged particles used four 21 x 21 x 130-cm 3 Bicron 408 scintillation counters to detect a signal at the single-photoelectron level. The competing noise requiring minimization was due to a combination of PM tube (8-inch Thorne EMI 9353KA) afterpulsing and ambient radiation-induced scintillator luminescence. A very slow decay (> 30 μs) component was observed and received particular attention. Efforts to reduce the SPE noise included photomultiplier tube base modifications, detector shielding and cooling, signal amplification, and veto procedures

  4. Radiation-hard semiconductor detectors for SuperLHC

    CERN Document Server

    Bruzzi, Mara; Al-Ajili, A A; Alexandrov, P; Alfieri, G; Allport, Philip P; Andreazza, A; Artuso, M; Assouak, S; Avset, B S; Barabash, L; Baranova, E; Barcz, A; Basile, A; Bates, R; Belova, N; Betta, G F D; Biagi, S F; Bilei, G M; Bisello, D; Blue, A; Blumenau, A; Boisvert, V; Bölla, G; Bondarenko, G B; Borchi, E; Borrello, L; Bortoletto, D; Boscardin, M; Bosisio, L; Bowcock, T J V; Brodbeck, T J; Broz, J; Brukhanov, A; Brzozowski, A; Buda, M; Buhmann, P; Buttar, C; Campabadal, F; Campbell, D; Candelori, A; Casse, G; Cavallini, A; Chilingarov, A G; Chren, D; Cindro, V; Citterio, M; Collins, P; Coluccia, R; Contarato, D; Coutinho, J; Creanza, D; Cunningham, W; Cvetkov, V; Davies, G; Dawson, I; De Palma, M; Demina, R; Dervan, P; Dierlamm, A; Dittongo, S; Dobrzanski, L; Dolezal, Z; Dolgolenko, A; Eberlein, T; Eremin, V; Fall, C; Fasolo, F; Ferbel, T; Fizzotti, F; Fleta, C; Focardi, E; Forton, E; Franchenko, S; Fretwurst, E; Gamaz, F; García-Navarro, J E; García, C; Gaubas, E; Genest, M H; Gill, K A; Giolo, K; Glaser, M; Gössling, C; Golovine, V; Gorelov, I; Goss, J; Gouldwell, A; Grégoire, G; Gregori, P; Grigoriev, E; Grigson, C; Grillo, A; Groza, A; Guskov, J; Haddad, L; Harding, R; Härkönen, J; Hauler, F; Hayama, S; Hoeferkamp, M; Honniger, F; Horazdovsky, T; Horisberger, R P; Horn, M; Houdayer, A; Hourahine, B; Hruban, A; Hughes, G; Ilyashenko, Yu S; Irmscher, K; Ivanov, A; Jarasiunas, K; Jin, T; Jones, B K; Jones, R; Joram, C; Jungermann, L; Kalinina, E; Kaminski, P; Karpenko, A; Karpov, A; Kazlauskiene, V; Kazukauskas, V; Khivrich, V; Khomenkov, V P; Kierstead, J A; Klaiber Lodewigs, J M; Kleverman, M; Klingenberg, R; Kodys, P; Kohout, Z; Korjenevski, S; Kowalik, A; Kozlowski, R; Kozodaev, M; Kramberger, G; Krasel, O; Kuznetsov, A; Kwan, S; Lagomarsino, S; Lari, T; Lassila-Perini, K M; Lastovetsky, V F; Latino, G; Latushkin, S T; Lazanu, I; Lazanu, S; Lebel, C; Leinonen, K; Leroy, C; Li, Z; Lindström, G; Lindström, L; Linhart, V; Litovchenko, A P; Litovchenko, P G; Litvinov, V; Lo Giudice, A; Lozano, M; Luczynski, Z; Luukka, Panja; Macchiolo, A; Mainwood, A; Makarenko, L F; Mandic, I; Manfredotti, C; Martí i García, S; Marunko, S; Mathieson, K; Melone, J; Menichelli, D; Meroni, C; Messineo, A; Miglio, S; Mikuz, M; Miyamoto, J; Moll, M; Monakhov, E; Moscatelli, F; Mozzanti, A; Murin, L; Naoumov, D; Nava, F; Nossarzhevska, E; Nummela, S; Nysten, J; Olivero, P; O'Shea, V; Palviainen, T; Paolini, C; Parkes, C; Passeri, D; Pein, U; Pellegrini, G; Perera, L; Petasecca, M; Piatkowski, B; Piemonte, C; Pignatel, G U; Pinho, N; Pintilie, I; Pintilie, L; Polivtsev, L; Polozov, P; Popa, A I; Popule, J; Pospísil, S; Pucker, G; Radicci, V; Rafí, J M; Ragusa, F; Rahman, M; Rando, R; Röder, R; Rohe, T; Ronchin, S; Rott, C; Roy, A; Roy, P; Ruzin, A; Ryazanov, A; Sadrozinski, H F W; Sakalauskas, S; Scaringella, M; Schiavulli, L; Schnetzer, S; Schumm, B; Sciortino, S; Scorzoni, A; Segneri, G; Seidel, S; Seiden, A; Sellberg, G; Sellin, P J; Sentenac, D; Sevilla, S G; Shipsey, I; Sícho, P; Sloan, T; Solar, M; Son, S; Sopko, B; Spencer, N; Stahl, J; Stavitski, I; Stolze, D; Stone, R; Storasta, J; Strokan, N; Strupinski, W; Sudzius, M; Surma, B; Suuronen, J; Suvorov, A; Svensson, B G; Tipton, P; Tomasek, M; Troncon, C; Tsvetkov, A; Tuominen, E; Tuovinen, E; Tuuva, T; Tylchin, M; Uebersee, H; Uher, J; Ullán, M; Vaitkus, J V; Vanni, P; Velthuis, J; Verbitskaya, E; Verzellesi, G; Vrba, V; Wagner, G; Wilhelm, I; Worm, S; Wright, V; Wunstorf, R; Zabierowski, P; Zaluzhny, A; Zavrtanik, M; Zen, M; Zhukov, V; Zorzi, N; de Boer, Wim

    2005-01-01

    An option of increasing the luminosity of the Large Hadron Collider (LHC) at CERN to 10/sup 35/ cm-/sup 2/s-/sup 1/ has been envisaged to extend the physics reach of the machine. An efficient tracking down to a few centimetres from the interaction point will be required to exploit the physics potential of the upgraded LHC. As a consequence, the semiconductor detectors close to the interaction region will receive severe doses of fast hadron irradiation and the inner tracker detectors will need to survive fast hadron fluences of up to above 10 /sup 16/ cm-/sup 2/. The CERN-RD50 project "Development of Radiation Hard Semiconductor Devices for Very High Luminosity Colliders" has been established in 2002 to explore detector materials and technologies that will allow to operate devices up to, or beyond, this limit. The strategies followed by RD50 to enhance the radiation tolerance include the development of new or defect engineered detector materials (SiC, GaN, Czochralski and epitaxial silicon, oxygen enriched Flo...

  5. X-ray radiation detectors of 'scintillator-photoreceiving device type' for industrial digital radiography with improved spatial resolution

    International Nuclear Information System (INIS)

    Ryzhykov, V.D.; Lysetska, O.K.; Opolonin, O.D.; Kozin, D.N.

    2003-01-01

    Main types of photo receivers used in X-ray digital radiography systems are luminescent screens that transfer the optical image onto charge collection instruments, which require cooling, and semiconductor silicon detectors, which limit the contrast sensitivity. We have developed and produced X-ray radiation detectors of 'scintillator-photoreceiving device' (S-PRD) type, which are integrally located on the inverse side of the photodiode (PD). The receiving-converting circuit (RCC) is designed for data conversion into digital form and their input into PC. Software is provided for RCC control and image visualization. Main advantages of these detectors are high industrial resolution (3-5 line pairs per mm), detecting activity up to 20 μm, controlled sensitivity, low weight and small size, imaging low (0.1-0.3 mrad) object dose in real time. In this work, main characteristics of 32-, 64- and 1024-channel detectors of S-PRD type were studied and compared for X-ray sensitivity with S-PD detectors. Images of the tested objects have been obtained. Recommendations are given on the use of different scintillation materials, depending upon the purpose of a digital radiographic system. The detectors operate in a broad energy range of ionizing radiation, hence the size of the controlled object is not limited. The system is sufficiently powerful to ensure frontal (through two walls) observation of pipelines with wall thickness up to 10 cm

  6. Hard electronics; Hard electronics

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1997-03-01

    Hard material technologies were surveyed to establish the hard electronic technology which offers superior characteristics under hard operational or environmental conditions as compared with conventional Si devices. The following technologies were separately surveyed: (1) The device and integration technologies of wide gap hard semiconductors such as SiC, diamond and nitride, (2) The technology of hard semiconductor devices for vacuum micro- electronics technology, and (3) The technology of hard new material devices for oxides. The formation technology of oxide thin films made remarkable progress after discovery of oxide superconductor materials, resulting in development of an atomic layer growth method and mist deposition method. This leading research is expected to solve such issues difficult to be easily realized by current Si technology as high-power, high-frequency and low-loss devices in power electronics, high temperature-proof and radiation-proof devices in ultimate electronics, and high-speed and dense- integrated devices in information electronics. 432 refs., 136 figs., 15 tabs.

  7. Scintillation response of organic and inorganic scintillators

    CERN Document Server

    Papadopoulos, L M

    1999-01-01

    A method to evaluate the scintillation response of organic and inorganic scintillators to different heavy ionizing particles is suggested. A function describing the rate of the energy consumed as fluorescence emission is derived, i.e., the differential response with respect to time. This function is then integrated for each ion and scintillator (anthracene, stilbene and CsI(Tl)) to determine scintillation response. The resulting scintillation responses are compared to the previously reported measured responses. Agreement to within 2.5% is observed when these data are normalized to each other. In addition, conclusions regarding the quenching parameter kB dependence on the type of the particle and the computed values of kB for certain ions are included. (author)

  8. Sub-aquatic response of a scintillator, fibre optic and silicon photomultiplier based radiation sensor

    Energy Technology Data Exchange (ETDEWEB)

    Jackson, Sarah F., E-mail: s.f.jackson@lancaster.ac.uk [Engineering Department, Lancaster University, Lancaster (United Kingdom); Monk, Stephen D., E-mail: s.monk@lancaster.ac.uk [Engineering Department, Lancaster University, Lancaster (United Kingdom); Stanley, Steven J., E-mail: steven.j.stanley@nnl.co.uk [National Nuclear Laboratory, A709 Springfields, Preston (United Kingdom); Lennox, Kathryn, E-mail: kathryn.lennox@nnl.co.uk [National Nuclear Laboratory, A709 Springfields, Preston (United Kingdom)

    2014-07-01

    We describe an attempt at the utilisation of two low level light sensors to improve on the design of a dose monitoring system, specifically for underwater applications with consideration for the effects of water attenuation. The gamma radiation ‘RadLine{sup ®}’ detector consists of an inorganic scintillating crystal coupled to a fibre optic cable which transports scintillation photons, up to hundreds of metres, to an optical sensor. Analysed here are two contemporary technologies; SensL's MiniSL a silicon photomultiplier (SiPM) and a Sens-Tech photon counting photomultiplier tube (PMT). A clinical radiotherapy linear accelerator (linac) is implemented as test beam, subjecting the RadLine{sup ®} to a highly controlled dose rate (ranging from 0 Sv h{sup −1} to 320 Sv h{sup −1}), averaging at 2 MeV in energy. The RadLine's underwater dose monitoring capabilities are tested with the aid of epoxy resin ‘solid water’ phantom blocks, used as a substitute for water. Our results show that the MiniSL SiPM is unsuitable for this application due to extremely high background noise levels, however the Sens-Tech PMT performs satisfactorily and the detected dose rate due to the effects of water attenuation compares strongly with MCNP simulation data and NIST database values. We conclude that the PMT shows promise for its ultimate use in the First Generation Magnox Storage Pond (FGMSP) on the Sellafield site. - Highlights: • RadLine{sup ®} consists of a scintillating crystal coupled to a fibre optic cable and photon detector. • Here the dose monitoring system is trialled with SiPM and PMT type photon detectors. • A clinical linear accelerator (linac) is used as a test beam. • Sub-aquatic response is compared to Monte Carlo simulations and the NIST database.

  9. Radiation hardness of β-Ga2O3 metal-oxide-semiconductor field-effect transistors against gamma-ray irradiation

    Science.gov (United States)

    Wong, Man Hoi; Takeyama, Akinori; Makino, Takahiro; Ohshima, Takeshi; Sasaki, Kohei; Kuramata, Akito; Yamakoshi, Shigenobu; Higashiwaki, Masataka

    2018-01-01

    The effects of ionizing radiation on β-Ga2O3 metal-oxide-semiconductor field-effect transistors (MOSFETs) were investigated. A gamma-ray tolerance as high as 1.6 MGy(SiO2) was demonstrated for the bulk Ga2O3 channel by virtue of weak radiation effects on the MOSFETs' output current and threshold voltage. The MOSFETs remained functional with insignificant hysteresis in their transfer characteristics after exposure to the maximum cumulative dose. Despite the intrinsic radiation hardness of Ga2O3, radiation-induced gate leakage and drain current dispersion ascribed respectively to dielectric damage and interface charge trapping were found to limit the overall radiation hardness of these devices.

  10. Scintillation properties of polycrystalline LaxY1-xO3 ceramic

    Science.gov (United States)

    Sahi, Sunil; Chen, Wei; Kenarangui, Rasool

    2015-03-01

    Scintillators are the material that absorbs the high-energy photons and emits visible photons. Scintillators are commonly used in radiation detector for security, medical imaging, industrial applications and high energy physics research. Two main types of scintillators are inorganic single crystals and organic (plastic or liquid) scintillators. Inorganic single crystals are expensive and difficult to grow in desire shape and size. Also, some efficient inorganic scintillator such as NaI and CsI are not environmental friendly. But on the other hand, organic scintillators have low density and hence poor energy resolution which limits their use in gamma spectroscopy. Polycrystalline ceramic can be a cost effective alternative to expensive inorganic single crystal scintillators. Here we have fabricated La0.2Y1.8O3 ceramic scintillator and studied their luminescence and scintillation properties. Ceramic scintillators were fabricated by vacuum sintering of La0.2Y1.8O3 nanoparticles at temperature below the melting point. La0.2Y1.8O3 ceramic were characterized structurally using XRD and TEM. Photoluminescence and radioluminescence studies were done using UV and X-ray as an excitation source. We have used gamma isotopes with different energy to studies the scintillation properties of La0.2Y1.8O3 scintillator. Preliminary studies of La0.2Y1.8O3 scintillator shows promising result with energy resolution comparable to that of NaI and CsI.

  11. Organic liquid scintillation detectors for on-the-fly neutron/gamma alarming and radionuclide identification in a pedestrian radiation portal monitor

    Science.gov (United States)

    Paff, Marc Gerrit; Ruch, Marc L.; Poitrasson-Riviere, Alexis; Sagadevan, Athena; Clarke, Shaun D.; Pozzi, Sara

    2015-07-01

    We present new experimental results from a radiation portal monitor based on the use of organic liquid scintillators. The system was tested as part of a 3He-free radiation portal monitor testing campaign at the European Commission's Joint Research Centre in Ispra, Italy, in February 2014. The radiation portal monitor was subjected to a wide range of test conditions described in ANSI N42.35, including a variety of gamma-ray sources and a 20,000 n/s 252Cf source. A false alarm test tested whether radiation portal monitors ever alarmed in the presence of only natural background. The University of Michigan Detection for Nuclear Nonproliferation Group's system triggered zero false alarms in 2739 trials. It consistently alarmed on a variety of gamma-ray sources travelling at 1.2 m/s at a 70 cm source to detector distance. The neutron source was detected at speeds up to 3 m/s and in configurations with up to 8 cm of high density polyethylene shielding. The success of on-the-fly radionuclide identification varied with the gamma-ray source measured as well as with which of two radionuclide identification methods was used. Both methods used a least squares comparison between the measured pulse height distributions to library spectra to pick the best match. The methods varied in how the pulse height distributions were modified prior to the least squares comparison. Correct identification rates were as high as 100% for highly enriched uranium, but as low as 50% for 241Am. Both radionuclide identification algorithms produced mixed results, but the concept of using liquid scintillation detectors for gamma-ray and neutron alarming in radiation portal monitor was validated.

  12. Scintillation properties of quantum-dot doped styrene based plastic scintillators

    International Nuclear Information System (INIS)

    Park, J.M.; Kim, H.J.; Hwang, Y.S.; Kim, D.H.; Park, H.W.

    2014-01-01

    We fabricated quantum-dot doped plastic scintillators in order to control the emission wavelength. We studied the characterization of the quantum-dots (CdSe/ZnS) and PPO (2, 5-diphenyloxazole) doped styrene based plastic scintillators. PPO is usually used as a dopant to enhance the scintillation properties of organic scintillators with a maximum emission wavelength of 380 nm. In order to study the scintillation properties of the quantum-dots doped plastic scintillators, the samples were irradiated with X-ray, photon, and 45 MeV proton beams. We observed that only PPO doped plastic scintillators shows a luminescence peak around 380 nm. However, both the quantum-dots and PPO doped plastic scintillators shows luminescence peaks around 380 nm and 520 nm. Addition of quantum-dots had shifted the luminescence spectrum from 380 nm (PPO) toward the region of 520 nm (Quantum-dots). Emissions with wavelength controllable plastic scintillators can be matched to various kinds of photosensors such as photomultiplier tubes, photo-diodes, avalanche photo-diodes, and CCDs, etc. Also quantum-dots doped plastic scintillator, which is irradiated 45 MeV proton beams, shows that the light yield of quantum-dots doped plastic scintillator is increases as quantum-dots doping concentration increases at 520 nm. And also the plastic scintillators were irradiated with Cs-137 γ-ray for measuring fluorescence decay time. -- Highlights: • Quantum-dot doped plastic scintillator is grown by the thermal polymerization method. • Quantum-dot doped plastic scintillators can control the emission wavelength to match with photo-sensor. • Quantum-dots and PPO doped plastic scintillators emitted luminescence peaks around 380 nm and 520 nm. • We observed the energy transfer from PPO to quantum-dot in the quantum-dot doped plastic scintillator

  13. Scintillation properties of quantum-dot doped styrene based plastic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Park, J.M.; Kim, H.J., E-mail: hongjooknu@gmail.com; Hwang, Y.S.; Kim, D.H.; Park, H.W.

    2014-02-15

    We fabricated quantum-dot doped plastic scintillators in order to control the emission wavelength. We studied the characterization of the quantum-dots (CdSe/ZnS) and PPO (2, 5-diphenyloxazole) doped styrene based plastic scintillators. PPO is usually used as a dopant to enhance the scintillation properties of organic scintillators with a maximum emission wavelength of 380 nm. In order to study the scintillation properties of the quantum-dots doped plastic scintillators, the samples were irradiated with X-ray, photon, and 45 MeV proton beams. We observed that only PPO doped plastic scintillators shows a luminescence peak around 380 nm. However, both the quantum-dots and PPO doped plastic scintillators shows luminescence peaks around 380 nm and 520 nm. Addition of quantum-dots had shifted the luminescence spectrum from 380 nm (PPO) toward the region of 520 nm (Quantum-dots). Emissions with wavelength controllable plastic scintillators can be matched to various kinds of photosensors such as photomultiplier tubes, photo-diodes, avalanche photo-diodes, and CCDs, etc. Also quantum-dots doped plastic scintillator, which is irradiated 45 MeV proton beams, shows that the light yield of quantum-dots doped plastic scintillator is increases as quantum-dots doping concentration increases at 520 nm. And also the plastic scintillators were irradiated with Cs-137 γ-ray for measuring fluorescence decay time. -- Highlights: • Quantum-dot doped plastic scintillator is grown by the thermal polymerization method. • Quantum-dot doped plastic scintillators can control the emission wavelength to match with photo-sensor. • Quantum-dots and PPO doped plastic scintillators emitted luminescence peaks around 380 nm and 520 nm. • We observed the energy transfer from PPO to quantum-dot in the quantum-dot doped plastic scintillator.

  14. Study of the Radiation-Hardness of VCSEL and PIN

    CERN Document Server

    Gan, K K; Fernando, W; Kagan, H P; Kass, R D; Lebbai, M R M; Merritt, H; Moore, J R; Nagarkar, A; Rizatdinova, F; Skubic, P L; Smith, D S; Strang, M

    2009-01-01

    The silicon trackers of the ATLAS experiment at the Large Hadron Collider (LHC) at CERN (Geneva) use optical links for data transmission. An upgrade of the trackers is planned for the Super LHC (SLHC), an upgraded LHC with ten times higher luminosity. We study the radiation-hardness of VCSELs (Vertical-Cavity Surface-Emitting Laser) and GaAs and silicon PINs using 24 GeV/c protons at CERN for possible application in the data transmission upgrade. The optical power of VCSEL arrays decreases significantly after the irradiation but can be partially annealed with high drive currents. The responsivities of the PIN diodes also decrease significantly after irradiation, but can be recovered by operating at higher bias voltage. This provides a simple mechanism to recover from the radiation damage.

  15. Scintillator structures

    International Nuclear Information System (INIS)

    Cusano, D.A.; Prener, J.S.

    1978-01-01

    Distributed phosphor scintillator structures providing superior optical coupling to photoelectrically responsive devices together with methods for fabricating said scintillator structures are disclosed. In accordance with one embodiment of the invention relating to scintillator structures, the phosphor is distributed in a 'layered' fashion with certain layers being optically transparent so that the visible wavelength output of the scintillator is better directed to detecting devices. In accordance with another embodiment of the invention relating to scintillator structures, the phosphor is distributed throughout a transparent matrix in a continuous fashion whereby emitted light is more readily transmitted to a photodetector. Methods for fabricating said distributed phosphor scintillator structures are also disclosed. (Auth.)

  16. Radiation hard silicon detectors - developments by the RD48(ROSE) collaboration

    Czech Academy of Sciences Publication Activity Database

    Lindström, G.; Kohout, Z.; Pospíšil, S.; Šícho, Petr; Sopko, B.; Vrba, Václav; Wilhelm, I.

    2001-01-01

    Roč. 466, č. 2 (2001), s. 308-326 ISSN 0168-9002 R&D Projects: GA MŠk LN00A006 Institutional research plan: CEZ:AV0Z1010920 Keywords : silicon detectors * radiation hardness * defect engineering * non ionizing energy los Subject RIV: BF - Elementary Particles and High Energy Physics Impact factor: 1.026, year: 2001

  17. Radiation hardness of silicon detectors manufactured on epitaxial material and FZ bulk enriched with oxygen, carbon, tin and platinum

    CERN Document Server

    Ruzin, A; Glaser, M; Lemeilleur, F; Talamonti, R; Watts, S; Zanet, A

    1999-01-01

    Recent results on the radiation hardness of silicon detectors fabricated on epitaxial and float zone bulk silicon enriched by various impurities, such as carbon, oxygen, tin and platinum are reported. A new methodology of measurements of electrical properties of the devices has been utilized in the experiment. It has been shown that in the case of irradiation by protons, oxygen enriched silicon has better radiation hardness than standard float zone silicon. The carbon enriched silicon detectors, on the other hand, exhibited significantly inferior radiation hardness compared to standard detectors. This study shows for the first time, a violation of the widely used normalization technique of the various particle irradiations by NIEL coefficients. The study has been carried out in the framework of the RD48 (ROSE) collaboration, which studies the radiation hardening of silicon detectors. (5 refs).

  18. Final Technical Report Radiation Hard Tight Pitch GaInP SPAD Arrays for High Energy Physics

    Energy Technology Data Exchange (ETDEWEB)

    Harmon, Eric

    2018-01-26

    The specialized photodetectors used in high energy physics experiments often need to remain extremely sensitive for years despite radiation induced damage caused by the constant bombardment of high energy particles. To solve this problem, LightSpin Technologies, Inc. in collaboration with Prof. Bradley Cox and the University of Virginia is developing radiation-hard GaInP photodetectors which are projected to be extraordinarily radiation hard, theoretically capable of withstanding a 100,000-fold higher radiation dose than silicon. In this Phase I SBIR project, LightSpin investigated the performance and radiation hardness of fifth generation GaInP SPAD arrays. These fifth generation devices used a new planar processing approach that enables very tight pitch arrays to be produced. High performance devices with SPAD pitches of 11, 15, and 25 μm were successfully demonstrated, which greatly increased the dynamic range and maximum count rate of the devices. High maximum count rates are critical when considering radiation hardness, since radiation damage causes a proportional increase in the dark count rate, causing SPAD arrays with low maximum count rates (large SPAD pitches) to fail. These GaInP SPAD array Photomultiplier Chips™ were irradiated with protons, electrons, and neutrons. Initial irradiation results were disappointing, with the post-irradiation devices exhibiting excessively high dark currents. The degradation was traced to surface leakage currents that were largely eliminated through the use of trenches etched around the exterior of the Photomultiplier Chip™ (not between SPAD elements). A second round of irradiations on Photomultiplier Chips™ with trenches proved substantially more successful, with post-irradiation dark currents remaining relatively low, though dark count rates were observed to increase at the highest doses. Preliminary analysis of the post-irradiation devices is promising … many of the irradiated Photomultiplier Chips™ still

  19. Liquid scintillation solutions

    International Nuclear Information System (INIS)

    Long, E.C.

    1976-01-01

    The liquid scintillation solution described includes a mixture of: a liquid scintillation solvent, a primary scintillation solute, a secondary scintillation solute, a variety of appreciably different surfactants, and a dissolving and transparency agent. The dissolving and transparency agent is tetrahydrofuran, a cyclic ether. The scintillation solvent is toluene. The primary scintillation solute is PPO, and the secondary scintillation solute is dimethyl POPOP. The variety of appreciably different surfactants is composed of isooctylphenol-polyethoxyethanol and sodium dihexyl sulphosuccinate [fr

  20. Comparison of proton microbeam and gamma irradiation for the radiation hardness testing of silicon PIN diodes

    Science.gov (United States)

    Jakšić, M.; Grilj, V.; Skukan, N.; Majer, M.; Jung, H. K.; Kim, J. Y.; Lee, N. H.

    2013-09-01

    Simple and cost-effective solutions using Si PIN diodes as detectors are presently utilized in various radiation-related applications in which excessive exposure to radiation degrades their charge transport properties. One of the conventional methods for the radiation hardness testing of such devices is time-consuming irradiation with electron beam or gamma-ray irradiation facilities, high-energy proton accelerators, or with neutrons from research reactors. Recently, for the purpose of radiation hardness testing, a much faster nuclear microprobe based approach utilizing proton irradiation has been developed. To compare the two different irradiation techniques, silicon PIN diodes have been irradiated with a Co-60 gamma radiation source and with a 6 MeV proton microbeam. The signal degradation in the silicon PIN diodes for both irradiation conditions has been probed by the IBIC (ion beam induced charge) technique, which can precisely monitor changes in charge collection efficiency. The results presented are reviewed on the basis of displacement damage calculations and NIEL (non-ionizing energy loss) concept.

  1. Radiation hardness studies for DEPFETs in Belle II

    International Nuclear Information System (INIS)

    Ritter, Andreas

    2014-01-01

    The study of CP violation requires dedicated detectors and accelerators. At KEK, the High Energy Accelerator Research Organization located in Tsukuba, Japan, an upgrade of the present accelerator KEKB and its detector is in progress. For this new Belle II detector, a new vertex system will be installed, consisting of a silicon strip detector (SVD) and a pixel detector (PXD). The PXD exhibits eight million pixels, each of them made of Depleted p-channel Field Effect Transistors (DEPFETs). During the operation of Belle II various machine- as well as luminosity-related background processes affect the device performance of the DEPFET through radiation damage. As a Metal-Oxide-Semiconductor (MOS) device, the DEPFET is affected by ionizing radiation damage as well as by damages to the silicon bulk itself. The major part of the radiation damage has its origin in the creation of electrons and positrons near the interaction point. Therefore, the hardness factor of electrons of relevant energy was investigated in this work. With this quantity the damage by electrons could be compared to the damage inflicted by neutrons. Neutron irradiations were performed with DEPFETs and related silicon material. The effects of leakage current increase and type inversion were studied. As the electron hardness investigation indicates, the bulk damage done to the DEPFET is small in comparison to the impact on the silicon dioxide layer of the device. Ionizing radiation results in a build-up of oxide charge, thus changing the device characteristics. Especially the threshold voltage of the DEPFET is shifted to more negative values. This shift has to be compensated during the operation of Belle II and is limited by device and system constraints, thus an overall small shift is desired. The changes in the device characteristics were investigated for the two gate electrodes of the DEPFET with respect to their biasing and production related issues. With an additional layer of silicon nitride and a

  2. Design of a scattering polarimeter for hard X-ray astronomy

    International Nuclear Information System (INIS)

    Costa, E.; Cinti, M.N.; Feroci, M.; Matt, G.; Rapisarda, M.

    1995-01-01

    The design of a new hard X-ray Compton scattering polarimeter based on scintillating fibre technology is presented and studied in detail by means of Monte Carlo calculations. Several different configurations and materials have been tested in order to optimise the sensitivity in the medium/high energy X-ray band. A high sensitivity over the energy band 20-200 keV is obtained for a two material configuration. The advantages deriving from employing a new scintillating material, the YAP (YAlO 3 ), are also discussed. (orig.)

  3. A gamma scintillation spectrometer

    Energy Technology Data Exchange (ETDEWEB)

    Symbalisty, S

    1952-07-01

    A scintillation type gamma ray spectrometer employing coincidence counting, designed and built at the Physics Department of the University of Western Ontario is described. The spectrometer is composed of two anthracene and photomultiplier radiation detectors, two pulse analyzing channels, a coincidence stage, three scalers and a high voltage stabilized supply. A preliminary experiment to test the operation of the spectrometer was performed and the results of this test are presented. (author)

  4. Scintillation response of CsI: Tl crystal under neutron, gamma, alpha particles and beta excitations

    Energy Technology Data Exchange (ETDEWEB)

    Pereira, Maria da Conceicao Costa; Madi Filho, Tufic; Lopes, Valdir Maciel; Berretta, Jose Roberto; Cardenas, Jose Patricio Nahuel, E-mail: macoper@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2015-07-01

    Among the converters of X and gamma radiation in light photons, known as scintillators, the one which is the most efficient emits photons with a wavelength near 400 nm. Particularly, among them, the cesium iodine doped with thallium (CsI:Tl) crystal is that which matches better between the light emission spectrum (peak at 540 nm) and the quantum sensitivity curve of the photodiodes and CCD (Charge Coupled Device). This explains the renewed interest in using this crystal as scintillator. Although the CsI:Tl crystal is commercially available, its local development would give the possibility to obtain it in different geometric configurations and coupling. Moreover, there is a special interest in studying new conditions that will alter the properties of this crystal in order to achieve a optimal level of its functional characteristics. Having an efficient national scintillator with low cost is a strategic opportunity to study the response of a detector applied to different types of radiation. The crystal of cesium iodide activated with thallium (CsI:Tl) has a high gamma detection efficiency per unit volume. In this paper, the CsI:Tl crystal, grown by the vertical Bridgman technique in evacuated silica ampoules and with the purpose of use as radiation detectors, is described. To evaluate the scintillator, measures of the thallium distribution in the crystal volume were taken, with overall efficiency score. The scintillator response was studied through gamma radiation from sources of {sup 137}Cs, {sup 60}Co, {sup 22}Na, {sup 54}Mn, {sup 131}I and {sup 99m}Tc; the beta radiation from source of {sup 90}Sr/{sup 90}Y, alpha particles from {sup 241}Am source and the scintillator response to neutrons from Am/Be source. The energetic resolution for {sup 137}Cs gamma rays (662 keV) was 10%. The results showed the validity of using the CsI:Tl crystal developed in our laboratory, in many applications in the area of radiation detectors. (author)

  5. Research in high energy physics: Scintillating fiber detector development for the SSC: Annual progress report

    International Nuclear Information System (INIS)

    Ruchti, R.C.

    1988-01-01

    The scintillating fiber detector development program at the University of Notre Dame is divided into several components. These include: Research on scintillating glass fiber materials; Research on scintillating plastic fiber materials; Research on scintillating liquids in fiber capillaries; Studies of improvements in image intensification and light amplification of appropriate test and development facilities at Notre Dame. The overall goal of the program is to develop efficient scintillating fiber detectors with long, optical attenuation length, and excellent radiation resistance properties for tracking and microvertex detectors and as component active sampling materials for scintillation calorimetry. We now discuss each of these programs in turn. 2 figs., 3 tabs

  6. Monitoring system for testing the radiation hardness of a KINTEX-7 FPGA

    Energy Technology Data Exchange (ETDEWEB)

    Cojocariu, L. N., E-mail: lucian.cojocariu@cern.ch [Horia Hulubei National Institute for R& D in Physics and Nuclear Engineering (Reactorului 30, MG, Romania) (Romania); Stefan cel Mare University of Suceava (Universitatii 13, Suceava, Romania) (Romania); Placinta, V. M., E-mail: vlad-mihai.placinta@cern.ch [Horia Hulubei National Institute for R& D in Physics and Nuclear Engineering (Reactorului 30, MG, Romania) (Romania); University POLITEHNICA of Bucharest (Splaiul Independentei 313, Bucharest, Romania) (Romania); Dumitru, L., E-mail: dlaur@nipne.ro [Horia Hulubei National Institute for R& D in Physics and Nuclear Engineering (Reactorului 30, MG, Romania) (Romania)

    2016-03-25

    A much more efficient Ring Imaging Cherenkov sub-detector system will be rebuilt in the second long shutdown of Large Hadron Collider for the LHCb experiment. Radiation-hard electronic components together with Commercial Off-The-Shelf ones will be used in the new Cherenkov photon detection system architecture. An irradiation program was foreseen to determine the radiation tolerance for the new electronic devices, including a Field Programmable Gate Array from KINTEX-7 family of XILINX. An automated test bench for online monitoring of the XC7K70T KINTEX-7 device operation in radiation conditions was designed and implemented by the LHCb Romanian group.

  7. Detectors for proton counting. Si-APD and scintillation detectors

    International Nuclear Information System (INIS)

    Kishimoto, Shunji

    2008-01-01

    Increased intensity of synchrotron radiation requests users to prepare photon pulse detectors having higher counting rates. As detectors for photon counting, silicon-avalanche photodiode (Si-APD) and scintillation detectors were chosen for the fifth series of detectors. Principle of photon detection by pulse and need of amplification function of the detector were described. Structure and working principle, high counting rate measurement system, bunch of electrons vs. counting rate, application example of NMR time spectroscopy measurement and comments for users were described for the Si-APD detector. Structure of scintillator and photomultiplier tube, characteristics of scintillator and performance of detector were shown for the NaI detector. Future development of photon pulse detectors was discussed. (T. Tanaka)

  8. Fluorescent compounds for plastic scintillation applications

    International Nuclear Information System (INIS)

    Pla-Dalmau, A.; Bross, A.D.

    1994-04-01

    Several 2-(2'-hydroxyphenyl)benzothiazole, -benzoxazole, and -benzimidazole derivatives have been prepared. Transmittance, fluorescence, light yield, and decay time characteristics of these compounds have been studied in a polystyrene matrix and evaluated for use in plastic scintillation detectors. Radiation damage studies utilizing a 60 C source have also been performed

  9. Process and devices of detection of hard electromagnetic or particle radiations using a superconducting element

    International Nuclear Information System (INIS)

    Drukier, A.K.; Valette, Claude; Waysand, Georges.

    1975-01-01

    The invention relates to processes and systems for the detection of hard electromagnetic or particle radiations and the sensors fitted to these systems. 'Hard radiations' means those whose energy is greater than a variable threshold, depending on the applications, but always more than 5 keV. The use of these sensors and the associated systems can therefore be envisaged in radiography and also in emission gammagraphy in the biological, anatomic and medical fields. In these processes, in order to detect a photon or a radiation particle, use is made of the transition phenomenon of a homogeneous grain of superconducting material of the first kind, from the metastable superconducting state to the normal state, under the effect of a photoelectron ejected by the impact of the photon or of the particle on the grain of superconducting material [fr

  10. Effects of plasma-deposited silicon nitride passivation on the radiation hardness of CMOS integrated circuits

    International Nuclear Information System (INIS)

    Clement, J.J.

    1980-01-01

    The use of plasma-deposited silicon nitride as a final passivation over metal-gate CMOS integrated circuits degrades the radiation hardness of these devices. The hardness degradation is manifested by increased radiation-induced threshold voltage shifts caused principally by the charging of new interface states and, to a lesser extent, by the trapping of holes created upon exposure to ionizing radiation. The threshold voltage shifts are a strong function of the deposition temperature, and show very little dependence on thickness for films deposited at 300 0 C. There is some correlation between the threshold voltage shifts and the hydrogen content of the PECVD silicon nitride films used as the final passivation layer as a function of deposition temperature. The mechanism by which the hydrogen contained in these films may react with the Si/SiO 2 interface is not clear at this point

  11. Crystal growth and characterization of europium doped lithium strontium iodide scintillator as an ionizing radiation detector

    Science.gov (United States)

    Uba, Samuel

    High performance detectors used in the detection of ionizing radiation is critical to nuclear nonproliferation applications and other radiation detectors applications. In this research we grew and tested Europium doped Lithium Strontium Iodide compound. A mixture of lithium iodide, strontium iodide and europium iodide was used as the starting materials for this research. Congruent melting and freezing temperature of the synthesized compound was determined by differential scanning calorimetry (DSC) using a Setaram Labsys Evo DSC-DTA instrument. The melting temperatures were recorded at 390.35°C, 407.59°C and freezing temperature was recorded at 322.84°C from a graph of heat flow plotted against temperature. The synthesized material was used as the charge for the vertical Bridgeman growth, and a 6.5 cm and 7.7cm length boule were grown in a multi-zone transparent Mullen furnace. A scintillating detector of thickness 2.53mm was fabricated by mechanical lapping in mineral oil, and scintillating response and timing were obtained to a cesium source using CS-137 isotope. An energy resolution (FWHM over peak position) of 12.1% was observed for the 662keV full absorption peak. Optical absorption in the UV-Vis wavelength range was recorded for the grown crystal using a U-2900 UV/VIS Spectrophotometer. Absorption peaks were recorded at 194nm, 273nm, and 344nm from the absorbance spectrum, various optical parameters such as absorption coefficient, extinction coefficient, refractive index, and optical loss were derived. The optical band gap energy was calculated using Tauc relation expression at 1.79eV.

  12. Gamma-ray detectors for intelligent, hand-held radiation monitors

    International Nuclear Information System (INIS)

    Fehlau, P.E.

    1983-01-01

    Small radiation detectors based on HgI 2 , bismuth germanate (BGO), plastic, or NaI(Tl) detector materials were evaluated for use in small, lighweight radiation monitors. The two denser materials, HgI 2 and BGO, had poor resolution at low-energy and thus performed less well than NaI(Tl) in detecting low-energy gamma rays from bare, enriched uranium. The plastic scintillator, a Compton recoil detector, also performed less well at low gamma-ray energy. Two small NaI(Tl) detectors were suitable for detecting bare uranium and sheilded plutonium. One became part of a new lightweight hand-held monitor and the other found uses as a pole-mounted detector for monitoring hard-to-reach locations

  13. Radiation hardness of CMOS monolithic active pixel sensors manufactured in a 0.18 μm CMOS process

    Energy Technology Data Exchange (ETDEWEB)

    Linnik, Benjamin [Goethe-Universitaet Frankfurt (Germany); Collaboration: CBM-MVD-Collaboration

    2015-07-01

    CMOS Monolithic Active Pixels Sensors (MAPS) are considered as the technology of choice for various vertex detectors in particle and heavy-ion physics including the STAR HFT, the upgrade of the ALICE ITS, the future ILC detectors and the CBM experiment at FAIR. To match the requirements of those detectors, their hardness to radiation is being improved, among others in a joined research activity of the Goethe University Frankfurt and the IPHC Strasbourg. It was assumed that combining an improved high resistivity (1-8 kΩcm) sensitive medium with the features of a 0.18 μm CMOS process, is suited to reach substantial improvements in terms of radiation hardness as compared to earlier sensor designs. This strategy was tested with a novel generation of sensor prototypes named MIMOSA-32 and MIMOSA-34. We show results on the radiation hardness of those sensors and discuss its impact on the design of future vertex detectors.

  14. Luminescent properties of composite scintillators based on PPO and o-POPOP doped SiO{sub 2} xerogel matrices

    Energy Technology Data Exchange (ETDEWEB)

    Viagin, O., E-mail: viagin@isma.kharkov.ua [Institute for Scintillation Materials of NAS of Ukraine, 60 Science Ave., 61001 Kharkiv (Ukraine); Masalov, A.; Bespalova, I.; Zelenskaya, O.; Tarasov, V.; Seminko, V.; Voloshina, L. [Institute for Scintillation Materials of NAS of Ukraine, 60 Science Ave., 61001 Kharkiv (Ukraine); Zorenko, Yu. [Institute of Physics of Kazimierz Wielki University of Bydgoszcz, 2 Powstańców Wielkopolskich str., 85-090 Bydgoszcz (Poland); Malyukin, Yu. [Institute for Scintillation Materials of NAS of Ukraine, 60 Science Ave., 61001 Kharkiv (Ukraine)

    2016-11-15

    New composite scintillation detectors were obtained by incorporation of PPO and o-POPOP organic scintillators into porous sol–gel silica matrices. Composites possess high photoluminescence intensity and decay time in nanosecond range. The absolute light yield of composite scintillators at excitation by alpha-radiation is about 4000–5000 photons/MeV and the pulse–height resolution is about 30%. The investigations of time-resolved luminescence of composites performed under excitation by synchrotron radiation in the 3.7–25 eV range have shown that the non-radiative energy transfer between host matrix and dopant molecules occurs via singlet states of SiO{sub 2} oxygen-deficient centers.

  15. Low-Dimensional Nanomaterials and Molecular Dielectrics for Radiation-Hard Electronics

    Science.gov (United States)

    McMorrow, Julian

    memory (SRAM) cells, an accomplishment that illustrates the technological relevance of this work by implementing a highly utilized component of modern day computing. Next, these SRAM devices demonstrate functionality as true random number generators (TRNGs), which are critical components in cryptography and encryption. The randomness of these SWCNT TRNGs is verified by a suite of statistical tests. This achievement has implications for securing data and communication in future solution-processed, large-area, flexible electronics. The unprecedented integration achieved by the underlying SWCNT doping and encapsulation motivates the study of this technology in a radiation environment. Doing so results in an understanding of the fundamental charge trapping mechanisms responsible for the radiation response in this system. The integrated nature of these devices enables, for the first time, the observation of system-level effects in a SWCNT integrated circuit technology. This technology is found to be total ionizing dose-hard, a promising result for the adoption of SWCNTs in future space-bound applications. Compared to SWCNTs, the field of MoS2 electronics is relatively nascent. As a result, studies of radiation effects in MoS2 devices focus on the fundamental mechanisms at play in the materials system. Here, we reveal the critical role of atmospheric adsorbates in the radiation effects of MoS2 transistors by measuring their response to vacuum ultraviolet radiation. These results highlight the importance of controlling the atmosphere of MoS2 devices during irradiation. Furthermore, we make recommendations for radiation-hard MoS2-based devices in the future as the technology continues to mature. One such recommendation is the incorporation of specialized dielectrics with proven radiation hardness. To this end, we address the materials integration challenge of incorporating SAND gate dielectrics on arbitrary substrates. We explore a novel approach for preparing metal substrates

  16. Scintillation properties of acrylate based plastic scintillator by photoploymerization method

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Sung Hwan [Dept. of Radiological Science, Cheongju University, Cheongju (Korea, Republic of); Lee, Joo Il [Dept. of of Radiology, Daegu Health College, Daegu (Korea, Republic of)

    2016-12-15

    In this study, we prepared and characterized a acrylate based UV-curable plastic scintillator. It was used co-polymers TMPTA, DHPA and Ultima GoldTM LLT organic scintillator. The emission spectrum of the plastic scintillator was located in the range of 380⁓520 nm, peaking at 423 nm. And the scintillator is more than 50% transparent in the range of 400⁓ 800 nm. The emission spectrum is well match to the quantum efficiency of photo-multiplier tube and the fast decay time of the scintillation is 12 ns, approximately. This scintillation material provides the possibility of combining 3D printing technology, and then the applications of the plastic scintillator may be expected in human dosimetry etc.

  17. Comparison among the scintillation and electrets techniques for emission of alpha radiation

    International Nuclear Information System (INIS)

    Moreno y Moreno, A.; Gammage, R.B.

    2005-01-01

    A comparison of the scintillation and of electret techniques for measurement of Uranium, Plutonium and transuranic elements in low level radioactivity is presented in this work. The samples of material to measure were obtained by the Research Department of the Oak Ridge National Laboratory (ORNL) of contaminated soils by radioactive material of plutonium and other transuranic elements. The samples in both techniques were measured to infinite thickness. The electret E - PERM of Rad Elec. Inc.; it was the available selected detector in 1993 and the measures with this technique were carried out by the Dr. R. B. Gammage and the Dr. K. E. Meyer of the Health and Safety Research Division of ORNL. The scintillation technique for measurement of the samples up mentioned it was carried out by the Dr. Augusto Moreno y Moreno in the facilities of DOSAR of ORNL. The comparative results of the two techniques agree just as it is shown in this work; but the scintillation technique offers the advantage of a considerable reduction of the exposure time and reading of the samples - electret, if it is compared with the immediate measurement of the scintillation technique that is described in this work. (Author)

  18. Radiation detection device and a radiation detection method

    International Nuclear Information System (INIS)

    Blum, A.

    1975-01-01

    A radiation detection device is described including at least one scintillator in the path of radiation emissions from a distributed radiation source; a plurality of photodetectors for viewing each scintillator; a signal processing means, a storage means, and a data processing means that are interconnected with one another and connected to said photodetectors; and display means connected to the data processing means to locate a plurality of radiation sources in said distributed radiation source and to provide an image of the distributed radiation sources. The storage means includes radiation emission response data and location data from a plurality of known locations for use by the data processing means to derive a more accurate image by comparison of radiation responses from known locations with radiation responses from unknown locations. (auth)

  19. High Resolution Tracking Devices Based on Capillaries Filled with Liquid Scintillator

    CERN Multimedia

    Bonekamper, D; Vassiltchenko, V; Wolff, T

    2002-01-01

    %RD46 %title\\\\ \\\\The aim of the project is to develop high resolution tracking devices based on thin glass capillary arrays filled with liquid scintillator. This technique provides high hit densities and a position resolution better than 20 $\\mu$m. Further, their radiation hardness makes them superior to other types of tracking devices with comparable performance. Therefore, the technique is attractive for inner tracking in collider experiments, microvertex devices, or active targets for short-lived particle detection. High integration levels in the read-out based on the use of multi-pixel photon detectors and the possibility of optical multiplexing allow to reduce considerably the number of output channels, and, thus, the cost for the detector.\\\\ \\\\New optoelectronic devices have been developed and tested: the megapixel Electron Bombarded CCD (EBCCD), a high resolution image-detector having an outstanding capability of single photo-electron detection; the Vacuum Image Pipeline (VIP), a high-speed gateable pi...

  20. Scintillation counter, maximum gamma aspect

    International Nuclear Information System (INIS)

    Thumim, A.D.

    1975-01-01

    A scintillation counter, particularly for counting gamma ray photons, includes a massive lead radiation shield surrounding a sample-receiving zone. The shield is disassembleable into a plurality of segments to allow facile installation and removal of a photomultiplier tube assembly, the segments being so constructed as to prevent straight-line access of external radiation through the shield into radiation-responsive areas. Provisions are made for accurately aligning the photomultiplier tube with respect to one or more sample-transmitting bores extending through the shield to the sample receiving zone. A sample elevator, used in transporting samples into the zone, is designed to provide a maximum gamma-receiving aspect to maximize the gamma detecting efficiency. (U.S.)

  1. Microfluidic Scintillation Detectors for High Energy Physics

    CERN Document Server

    Maoddi, Pietro; Mapelli, Alessandro

    This thesis deals with the development and study of microfluidic scintillation detectors, a technology of recent introduction for the detection of high energy particles. Most of the interest for such devices comes from the use of a liquid scintillator, which entails the possibility of changing the active material in the detector, leading to increased radiation resistance. A first part of the thesis focuses on the work performed in terms of design and modelling studies of novel prototype devices, hinting to new possibilities and applications. In this framework, the simulations performed to validate selected designs and the main technological choices made in view of their fabrication are addressed. The second part of this thesis deals with the microfabrication of several prototype devices. Two different materials were studied for the manufacturing of microfluidic scintillation detectors, namely the SU-8 photosensitive epoxy and monocrystalline silicon. For what concerns the former, an original fabrication appro...

  2. Radiation hardness of GaAs sensors against gamma-rays, neutrons and electrons

    Energy Technology Data Exchange (ETDEWEB)

    Šagátová, Andrea, E-mail: andrea.sagatova@stuba.sk [Institute of Nuclear and Physical Engineering, Faculty of Electrical Engineering and Information Technology, Slovak University of Technology, Ilkovičova 3, 812 19 Bratislava (Slovakia); University Centre of Electron Accelerators, Slovak Medical University, Ku kyselke 497, 911 06 Trenčín (Slovakia); Zaťko, Bohumír; Dubecký, František [Institute of Electrical Engineering, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava (Slovakia); Ly Anh, Tu [Faculty of Applied Science, University of Technology VNU HCM, 268 Ly Thuong Kiet Street, District 10, Ho Chi Minh City (Viet Nam); Nečas, Vladimír; Sedlačková, Katarína; Pavlovič, Márius [Institute of Nuclear and Physical Engineering, Faculty of Electrical Engineering and Information Technology, Slovak University of Technology, Ilkovičova 3, 812 19 Bratislava (Slovakia); Fülöp, Marko [University Centre of Electron Accelerators, Slovak Medical University, Ku kyselke 497, 911 06 Trenčín (Slovakia)

    2017-02-15

    Highlights: • Radiation hardness of SI GaAs detectors against gamma-rays, neutrons and electrons was compared. • Good agreement was achieved between the experimental results and displacement damage factor of different types of radiation. • CCE and FWHM first slightly improved (by 1–8%) and just then degraded with the cumulative dose. • An increase of detection efficiency with cumulative dose was observed. - Abstract: Radiation hardness of semi-insulating GaAs detectors against {sup 60}Co gamma-rays, fast neutrons and 5 MeV electrons was compared. Slight improvements in charge collection efficiency (CCE) and energy resolution in FWHM (Full Width at Half Maximum) were observed at low doses with all kinds of radiation followed by their degradation. The effect occurred at a dose of about 10 Gy of neutrons (CCE improved by 1%, FWHM by 5% on average), at 1 kGy of electrons (FWHM decreased by 3% on average) and at 10 kGy of gamma-rays (CCE raised by 5% and FWHM dropped by 8% on average), which is in agreement with the relative displacement damage of the used types of radiation. Gamma-rays of MeV energies are 1000-times less damaging than similar neutrons and electrons about 10-times more damaging than photons. On irradiating the detectors with neutrons and electrons, we observed a global increase in their detection efficiency, which was caused probably by enlargement of the active detector area as a consequence of created radiation defects in the base material. Detectors were still functional after a dose of 1140 kGy of ∼1 MeV photons, 104 kGy of 5 MeV electrons but only up to 0.576 kGy of fast (∼2 to 30 MeV) neutrons.

  3. Radiation detection systems

    International Nuclear Information System (INIS)

    Rozsa, Sandor

    1989-01-01

    Principles and operation of radiation detectors used for industrial and laboratory radiometric measurements are reviewed. The discussed detector types are as follows: ionization chamber, proportional counter, Geiger-Mueller counter, semiconductor detectors, scintillation counter and scintillators, photomultiplier tube, scintillation spectrometer and scintillation detector with semiconductor light sensor. A brief overview of the detectors of industrial nuclear instruments is also presented. (R.P.) 32 figs.; 3 tabs

  4. Progress report for the scintillator plate calorimeter subsystem

    International Nuclear Information System (INIS)

    1990-01-01

    This report covers the work completed in FY90 by ANL staff and those of Westinghouse STC and BICRON Corporation under subcontract to ANL towards the design of a compensating calorimeter based on the use of scintillator plate as the sensitive medium. It is presented as five task sections dealing with respectively mechanical design; simulation studies; optical system design; electronics development; development of rad hard plastic scintillator and wavelength shifter and a summary. The work carried out by the University of Tennessee under a subcontract from ANL is reported separately. Finally, as principal institution with responsibility for the overall management of this subsystem effort, the summary here reports the conclusions resulting from the work of the collaboration and their impact on our proposed direction of effort in FY91. This proposal, for obvious reasons is given separately

  5. Radiation hardness of superconducting magnet insulation materials for FAIR

    International Nuclear Information System (INIS)

    Seidl, Tim

    2013-03-01

    This thesis focuses on radiation degradation studies of polyimide, polyepoxy/glass-fiber composites and other technical components used, for example, in the superconducting magnets of new ion accelerators such as the planned International Facility for Antiproton and Ion Research (FAIR) at the GSI Helmholtz Center of Heavy Ion Research (GSI) in Darmstadt. As accelerators are becoming more powerful, i.e., providing larger energies and beam intensities, the potential risk of radiation damage to the components increases. Reliable data of the radiation hardness of accelerator materials and components concerning electrical, thermal and other technical relevant properties are of great interest also for other facilities such as the Large Hadron Collider (LHC) of CERN. Dependent on the position of the different components, induced radiation due to beam losses consists of a cocktail of gammas, neutrons, protons, and heavier particles. Although the number of heavy fragments of the initial projectiles is small compared to neutrons, protons, or light fragments (e.g. ? particles), their large energy deposition can induce extensive damage at rather low fluences (dose calculations show that the contribution of heavy ions to the total accumulated dose can reach 80 %). For this reason, defined radiation experiments were conducted using different energetic ion beams (from protons to uranium) and gamma radiation from a Co-60 source. The induced changes were analyzed by means of in-situ and ex-situ analytical methods, e.g. ultraviolet-visible and infrared spectroscopy, residual gas analysis, thermal gravimetric analysis, dielectric strength measurements, measurements of low temperature thermal properties, and performance tests. In all cases, the radiation induces a change in molecular structure as well as loss of functional material properties. The amount of radiation damage is found to be sensitive to the used type of ionizing radiation and the long term stability of the materials is

  6. X-ray detection capabilities of plastic scintillators incorporated with hafnium oxide nanoparticles surface-modified with phenyl propionic acid

    Science.gov (United States)

    Hiyama, Fumiyuki; Noguchi, Takio; Koshimizu, Masanori; Kishimoto, Shunji; Haruki, Rie; Nishikido, Fumihiko; Yanagida, Takayuki; Fujimoto, Yutaka; Aida, Tsutomu; Takami, Seiichi; Adschiri, Tadafumi; Asai, Keisuke

    2018-01-01

    We synthesized plastic scintillators incorporated with HfO2 nanoparticles as detectors for X-ray synchrotron radiation. Nanoparticles with sizes of less than 10 nm were synthesized with the subcritical hydrothermal method. The detection efficiency of high-energy X-ray photons improved by up to 3.3 times because of the addition of the nanoparticles. Nanosecond time resolution was successfully achieved for all the scintillators. These results indicate that this method is applicable for the preparation of plastic scintillators to detect X-ray synchrotron radiation.

  7. Radiation imaging

    Energy Technology Data Exchange (ETDEWEB)

    Redmayne, I.

    1986-05-21

    A detector for the detection of radiation such as X-ray radiation comprises an array of scintillation elements embedded in a sheet of radiation absorbing material. The scintillation elements are monitored individually, for example by a corresponding array of photodiodes, to build up a picture of the incident radiation. The front face of the sheet and the inner walls of the bores may be coated with a reflective material. The detector finds particular application in weld radiography. The detector may be stepped relative to the radiation source, the signals produced by the rows of the detector as they pass a predetermined point being summed.

  8. Radiation imaging

    International Nuclear Information System (INIS)

    Redmayne, Ian.

    1986-01-01

    A detector for the detection of radiation such as X-ray radiation comprises an array of scintillation elements embedded in a sheet of radiation absorbing material. The scintillation elements are monitored individually, for example by a corresponding array of photodiodes, to build up a picture of the incident radiation. The front face of the sheet and the inner walls of the bores may be coated with a reflective material. The detector finds particular application in weld radiography. The detector may be stepped relative to the radiation source, the signals produced by the rows of the detector as they pass a predetermined point being summed. (author)

  9. Characterization of plastic scintillators for detection of radioactivity: Light yield, Time decay measurements and Neutron/γ Pulse Shape Discrimination

    International Nuclear Information System (INIS)

    Montbarbon, E.; Pansu, R.B.; Hamel, M.; Coulon, R.

    2015-07-01

    Since Helium-3 shortage, organic scintillators play a major role in neutron detection. CEA LIST decided to focus on plastic scintillators. By definition, a plastic scintillator is a radio-luminescent polymer; this means that it emits light after interaction with an ionizing radiation. A platform was developed to characterize lab-made prepared scintillators and to compare them with commercial scintillators. Three physicochemical criteria are determined with this unique platform. (authors)

  10. A miniature gamma ray dosimeter with CWO scintillator and plastic optical fiber combination

    International Nuclear Information System (INIS)

    Jae Woo Park, Min Woo Seo

    2008-01-01

    Full text: Fiber-optic scintillation dosimeters possess several favorable characteristics, such as remote measurability and superior spatial resolution. Such a radiation dosimeter model was developed by attaching a small piece of CWO (CdWO 4 ) scintillator to a low attenuation plastic optical fiber. CWO was chosen since the higher atomic numbers of Cd and W would render the size of the scintillator smaller. The size of the scintillator was 4.7x4.7x10 mm 3 . The scintillator was optically glued to the plastic optical fiber of 3 mm diameter and 10 m length. A current-type PMT was optically coupled to the other end of the fiber to convert the lights generated in the scintillator into current signals. The dosimeter model was tested with two 60 Co standard sources of 0.5 mCi and 1 mCi to measure the PMT current as a function of the source-to-detector distance. It was then tested in a 60 Co irradiation chamber with an activity of about 6600 Ci. MCNPX simulations were performed for the source and detector arrangements to calculate the deposited energy in the CWO scintillator. The profiles of the measured current change are compared with those of the calculated energy deposition change. While there is some deviation between the measured and calculated profiles obtained with the lower-activity standard sources, the measured profile accurately coincides with the calculated one obtained in the higher-activity irradiation chamber. This study suggests that the fiber-optic scintillation dosimeter, operated in current mode, can be used to remotely measure radiation doses in high-intensity gamma fields

  11. Photoluminescence and radiation response properties of Ce3+-doped CsCaCl3 crystalline scintillator

    International Nuclear Information System (INIS)

    Fujimoto, Yutaka; Saeki, Keiichiro; Tanaka, Hironori; Yahaba, Takuma; Koshimizu, Masanori; Asai, Keisuke; Yanagida, Takayuki

    2016-01-01

    In this paper, we report on the photoluminescence and scintillation properties of a newly developed CsCaCl 3 :Ce (0.5 mol%) crystalline scintillator grown by the vertical Bridgman method. The fluorescence quantum efficiency for the Ce 3+ characteristic emission bands centered at around 350–400 nm was 76% under excitation at 330 nm light. The photoluminescence decay time of the Ce 3+ was approximately 32 ns. When x-ray excited the crystal, intense emission bands were observed at 350–400 nm, and could be attributed to the Ce 3+ emission. The scintillation light yield of the developed crystal was ∼7600 ph MeV −1 compared to a NaI:Tl commercial scintillator, and the principal scintillation decay time was approximately 340 ns plus two fast components of around 1.6 ns and 45 ns. (paper)

  12. Recipe for attaining optimal energy resolution in inorganic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Singh, Jai; Koblov, Alexander [School of Engineering and IT, B-purple-12, Faculty of EHSE, Charles Darwin University, Darwin, NT 0909 (Australia)

    2012-12-15

    Using an approximate form of the density of excitation created within the track initiated by an incident {gamma} - photon on a scintillator, the light yield is derived as a function of linear, bimolecular and Auger radiative and quenching recombination rates. The non-proportionality in the yield is analysed as a function of the bimolecular and Auger quenching rates and also its dependence on the track radius is studied. An optimal combination of these quenching rates and track radius is presented to obtain a recipe for inventing a scintillator material with optimal energy resolution. The importance of the mobility of charge carriers in minimising the non-proportionality in a scintillator is also discussed (copyright 2012 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  13. Recipe for attaining optimal energy resolution in inorganic scintillators

    International Nuclear Information System (INIS)

    Singh, Jai; Koblov, Alexander

    2012-01-01

    Using an approximate form of the density of excitation created within the track initiated by an incident γ - photon on a scintillator, the light yield is derived as a function of linear, bimolecular and Auger radiative and quenching recombination rates. The non-proportionality in the yield is analysed as a function of the bimolecular and Auger quenching rates and also its dependence on the track radius is studied. An optimal combination of these quenching rates and track radius is presented to obtain a recipe for inventing a scintillator material with optimal energy resolution. The importance of the mobility of charge carriers in minimising the non-proportionality in a scintillator is also discussed (copyright 2012 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  14. Spatially distributed scintillator arrays for diagnosing runaway electron transport and energy behavior in tokamaks

    International Nuclear Information System (INIS)

    James, A. N.; Hollmann, E. M.; Tynan, G. R.

    2010-01-01

    We present details of a new bismuth germanate [Bi 4 Ge 3 O 12 (BGO)] scintillator array used to diagnose the transport and energy behavior of runaway electrons (REs) in DIII-D. BGO exhibits important properties for these compact detectors including high light yield which sufficiently excites photodiode detectors (8500 photons/MeV), high density and atomic numbers of constituent materials which maximizes sensitivity, and relative neutron blindness which minimizes complications in data interpretation. The detectors observe primarily hard x-ray radiation emitted in a forward beamed pattern by RE when they strike first wall materials or bulk ions and neutrals in the plasma, although we also address photoneutron signals. The arrangement of the array enables time resolved location of x-ray emission and associated asymmetries which help identify instabilities and confinement properties of RE. By shielding a subset of detectors with different thicknesses of lead, and with interpretative support of the code EGSNRC, we also measure RE energy, although due to the often distributed nature of RE strike points and the forward beamed character of emitted hard x-rays, we restrict interpretation as a lower bound for RE energy.

  15. Spatially distributed scintillator arrays for diagnosing runaway electron transport and energy behavior in tokamaks

    Energy Technology Data Exchange (ETDEWEB)

    James, A. N.; Hollmann, E. M.; Tynan, G. R. [UC San Diego Center for Energy Research, La Jolla, California 92093-0417 (United States)

    2010-10-15

    We present details of a new bismuth germanate [Bi{sub 4}Ge{sub 3}O{sub 12} (BGO)] scintillator array used to diagnose the transport and energy behavior of runaway electrons (REs) in DIII-D. BGO exhibits important properties for these compact detectors including high light yield which sufficiently excites photodiode detectors (8500 photons/MeV), high density and atomic numbers of constituent materials which maximizes sensitivity, and relative neutron blindness which minimizes complications in data interpretation. The detectors observe primarily hard x-ray radiation emitted in a forward beamed pattern by RE when they strike first wall materials or bulk ions and neutrals in the plasma, although we also address photoneutron signals. The arrangement of the array enables time resolved location of x-ray emission and associated asymmetries which help identify instabilities and confinement properties of RE. By shielding a subset of detectors with different thicknesses of lead, and with interpretative support of the code EGSNRC, we also measure RE energy, although due to the often distributed nature of RE strike points and the forward beamed character of emitted hard x-rays, we restrict interpretation as a lower bound for RE energy.

  16. Collimator trans-axial tomographic scintillation camera

    International Nuclear Information System (INIS)

    Jaszczak, R.J.

    1977-01-01

    A collimator is provided for a scintillation camera system in which a detector precesses in an orbit about a patient. The collimator is designed to have high resolution and lower sensitivity with respect to radiation traveling in paths laying wholly within planes perpendicular to the cranial-caudal axis of the patient. The collimator has high sensitivity and lower resolution to radiation traveling in other planes. Variances in resolution and sensitivity are achieved by altering the length, spacing or thickness of the septa of the collimator

  17. Response function study of a scintillator detector of NaI(Tl)

    International Nuclear Information System (INIS)

    Villa, Marcelo Barros; Costa, Alessandro Martins da

    2014-01-01

    In measurements of gamma rays with Nai (Tl) scintillator, the detectors output data are pulse height spectra, that corresponding to distorted information about the radiation source due to various errors associated with the crystal scintillation process and electronics associated, instead of power spectra photons. Pulse height spectra are related to the real power spectra by means of scintillator detector response function NaI (Tl). In this work, the response function for a cylindrical crystal of Nal (Tl) of 7,62 x 7,62 cm (diameter x length) was studied, by Monte Carlo method, using the EGSnrc tool to model the transport of radiation, combined with experimental measurements. An inverse response matrix, even with the energy of the square root, which transforms the pulse height spectrum of photon energy spectrum was obtained. The results of this transformation of pulse height spectrum for photon energy spectrum is presented, showing that the methodology employed in this study is suitable

  18. Radiation hardness of CMS pixel barrel modules

    International Nuclear Information System (INIS)

    Rohe, T.; Bean, A.; Erdmann, W.; Kaestli, H.-C.; Khalatyan, S.; Meier, B.; Radicci, V.; Sibille, J.

    2010-01-01

    Pixel detectors are used in the innermost part of the multi purpose experiments at the LHC and are therefore exposed to the highest fluences of ionising radiation, which in this part of the detectors consists mainly of charged pions. The radiation hardness of all detector components has been thoroughly tested up to the fluences expected at the LHC. In case of an LHC upgrade, the fluence will be much higher and it is not yet clear how long the present pixel modules will stay operative in such a harsh environment. The aim of this study was to establish such a limit as a benchmark for other possible detector concepts considered for the upgrade. As the sensors and the readout chip are the parts most sensitive to radiation damage, samples consisting of a small pixel sensor bump-bonded to a CMS-readout chip (PSI46V2.1) have been irradiated with positive 200 MeV pions at PSI up to 6x10 14 n eq /cm 2 and with 21 GeV protons at CERN up to 5x10 15 n eq /cm 2 . After irradiation the response of the system to beta particles from a 90 Sr source was measured to characterise the charge collection efficiency of the sensor. Radiation induced changes in the readout chip were also measured. The results show that the present pixel modules can be expected to be still operational after a fluence of 2.8x10 15 n eq /cm 2 . Samples irradiated up to 5x10 15 n eq /cm 2 still see the beta particles. However, further tests are needed to confirm whether a stable operation with high particle detection efficiency is possible after such a high fluence.

  19. Polyethylene Naphthalate Scintillator: A Novel Detector for the Dosimetry of Radioactive Ophthalmic Applicators.

    Science.gov (United States)

    Flühs, Dirk; Flühs, Andrea; Ebenau, Melanie; Eichmann, Marion

    2015-09-01

    Dosimetric measurements in small radiation fields with large gradients, such as eye plaque dosimetry with β or low-energy photon emitters, require dosimetrically almost water-equivalent detectors with volumes of <1 mm(3) and linear responses over several orders of magnitude. Polyvinyltoluene-based scintillators fulfil these conditions. Hence, they are a standard for such applications. However, they show disadvantages with regard to certain material properties and their dosimetric behaviour towards low-energy photons. Polyethylene naphthalate, recently recognized as a scintillator, offers chemical, physical and basic dosimetric properties superior to polyvinyltoluene. Its general applicability as a clinical dosimeter, however, has not been shown yet. To prove this applicability, extensive measurements at several clinical photon and electron radiation sources, ranging from ophthalmic plaques to a linear accelerator, were performed. For all radiation qualities under investigation, covering a wide range of dose rates, a linearity of the detector response to the dose was shown. Polyethylene naphthalate proved to be a suitable detector material for the dosimetry of ophthalmic plaques, including low-energy photon emitters and other small radiation fields. Due to superior properties, it has the potential to replace polyvinyltoluene as the standard scintillator for such applications.

  20. Radiation hard silicon microstrip detectors for Tevatron experiments

    International Nuclear Information System (INIS)

    Korjenevski, Sergey

    2004-01-01

    The Silicon Microstrip Tracking detectors at the CDF and D0 experiments have now been operating for almost three years at Fermilab. These detectors were designed originally for an integrated luminosity of 2fb -1 . As the expected luminosity for Run IIb at the Tevatron collider was initially envisioned to reach 15fb -1 , radiation tolerances of both devices were revisited, culminating in proposals for new systems. With reduced expectations for total luminosity at ∼6fb -1 , the full detector-replacement projects were terminated. The CDF detector is expected nevertheless to cope efficiently with the lower anticipated dose, however, the D0 experiment is planning a smaller-scale project: a Layer-0 (L0) upgrade of the silicon tracker (D0SMT). The new device will fit between the beam line and the inner layer of the current Tracker. Built of single-sided sensors, this upgrade is expected to perform well in the harsh radiation environment, and be able to withstand an integrated luminosity of 15fb -1 . Prototypes of Run IIb sensors were irradiated using 10MeV protons at the tandem Van de Graaff at the James R. McDonald Laboratory at Kansas State University. A fit to the 10MeV proton data yields a damage parameter αp=11x10-17Acm. This is consistent with results from RD48 (αp=9.9x10-17Acm). The scaling of damage to 1MeV neutron fluence uses a hardness factor (κ) derived from the non-ionizing components of the energy loss (NEIL). NEIL predicts a hardness factor of 3.87 for 10MeV protons. We obtained an experimental value of this factor of 2.54, or 34% smaller than scaling predictions from NEIL

  1. Development of crystals based in cesium iodide for application as radiation detectors; Desenvolvimento de cristais baseados em iodeto de cesio para aplicacao como detectores de radiacao

    Energy Technology Data Exchange (ETDEWEB)

    Pereira, Maria da Conceicao Costa

    2006-07-01

    Inorganic scintillators with fast luminescence decay time, high density and high light output have been the object of studies for application in nuclear physics, high energy physics, nuclear tomography and other fields of science and engineering. Scintillation crystals based on cesium iodide (CsI) are matters with relatively low higroscopy, high atomic number, easy handling and low cost, characteristics that favor their use as radiation detectors. In this work, the growth of pure CsI crystals, CsI:Br and CsI:Pb, using the Bridgman technique, is described. The concentration of the bromine doping element (Br) was studied in the range of 1,5x10{sup -1} M to 10{sup -2} M and the lead (Pb) in the range of 10{sup -2} M to 5x10{sup -4} M. To evaluate the scintillators developed, systematic measurements were carried out for luminescence emission and luminescence decay time for gamma radiation, optical transmittance assays, Vickers micro-hardness assays, determination of the doping elements distribution along the grown crystals and analysis of crystals response to the gamma radiation in the energy range of 350 keV to 1330 keV and alpha particles from a {sup 241}Am source, with energy of 5.54 MeV. It was obtained 13 ns to 19 ns for luminescence decay time for CsI:Br and CsI:Pb crystals. These results were very promising. The results obtained for micro-hardness showed a significant increase in function of the doping elements concentration, when compared to the pure CsI crystal, increasing consequently the mechanical resistance of the grown crystals. The validity of using these crystals as radiation sensors may be seen from the results of their response to gamma radiation and alpha particles. (author)

  2. Scintillation detectors in experiments on plasma accelerators

    International Nuclear Information System (INIS)

    Bystritskij, V.M.; Gerasimov, V.V.; Kublikov, R.V.; Parzhitskij, S.S.; Smirnov, V.S.; Wozniak, J.; Dudkin, G.N.; Nechaev, B.A.; Padalko, V.M.

    2005-01-01

    The gating circuits for photomultipliers of scintillation detectors operating in powerful pulsed electromagnetic and nuclear radiation fields are investigated. PMTs with the jalousie-type dynode system and with the linear dynode system are considered. The basic gating circuits of the photomultipliers involving active and resistor high-voltage dividers are given. The results of the investigations are important for experiments in which it is necessary to discriminate in time the preceding background radiation and the process of interest. (author)

  3. Validated Models for Radiation Response and Signal Generation in Scintillators: Final Report

    Energy Technology Data Exchange (ETDEWEB)

    Kerisit, Sebastien N.; Gao, Fei; Xie, YuLong; Campbell, Luke W.; Van Ginhoven, Renee M.; Wang, Zhiguo; Prange, Micah P.; Wu, Dangxin

    2014-12-01

    This Final Report presents work carried out at Pacific Northwest National Laboratory (PNNL) under the project entitled “Validated Models for Radiation Response and Signal Generation in Scintillators” (Project number: PL10-Scin-theor-PD2Jf) and led by Drs. Fei Gao and Sebastien N. Kerisit. This project was divided into four tasks: 1) Electronic response functions (ab initio data model) 2) Electron-hole yield, variance, and spatial distribution 3) Ab initio calculations of information carrier properties 4) Transport of electron-hole pairs and scintillation efficiency Detailed information on the results obtained in each of the four tasks is provided in this Final Report. Furthermore, published peer-reviewed articles based on the work carried under this project are included in Appendix. This work was supported by the National Nuclear Security Administration, Office of Nuclear Nonproliferation Research and Development (DNN R&D/NA-22), of the U.S. Department of Energy (DOE).

  4. Feasibility Study on Fiber-optic Radiation Sensor for Remote Gamma-ray Spectroscopy

    International Nuclear Information System (INIS)

    Jeon, Hyesu; Jang, Kyoung Won; Shin, Sang Hun and others

    2014-01-01

    In this study, we fabricated a fiber-optic radiation sensor using an optical fiber and various scintillators. To select an adequate inorganic scintillator for the sensing probe of fiber-optic radiation sensor, 5 types of scintillators were evaluated. The spectra of gamma-rays emitted from a Na-22 radiation source were measured by using the manufactured sensors. As a result, the BGO was suitable for the sensing probe of fiber-optic radiation sensor due to its high scintillation output and exact photoelectric peak for the gamma-ray energy. The basic principle of radiation detection is to detect the signals caused by interactions between radiations and materials. There are various types of radiation detectors depending on types of radiation to be detected and physical quantities to be measured. As one of the radiation detectors, a fiber-optic radiation sensor using a scintillator and an optical fiber has two advantages such as no space restraint and remote sensing. Moreover, in nuclear environments, this kind of sensor has immunities for electromagnetic field, temperature, and pressure. Thus, the fiber-optic radiation sensor can be used in various fields including nondestructive inspection, radioactive waste management, nuclear safety, radiodiagnosis and radiation therapy. As a fundamental study of the fiber-optic radiation sensor for remote gamma-ray spectroscopy, in this study, we fabricated a fiber-optic radiation sensor using an optical fiber and various scintillators. To select an adequate inorganic scintillator for the sensing probe of fiber-optic radiation sensor, 5 types of scintillators were evaluated. The spectra of gamma-rays emitted from a Na-22 radiation source were measured by using the manufactured sensors

  5. Novel scintillators and silicon photomultipliers for nuclear physics and applications

    International Nuclear Information System (INIS)

    Jenkins, David

    2015-01-01

    Until comparatively recently, scintillator detectors were seen as an old-fashioned tool of nuclear physics with more attention being given to areas such as gamma-ray tracking using high-purity germanium detectors. Next-generation scintillator detectors, such as lanthanum bromide, which were developed for the demands of space science and gamma- ray telescopes, are found to have strong applicability to low energy nuclear physics. Their excellent timing resolution makes them very suitable for fast timing measurements and their much improved energy resolution compared to conventional scintillators promises to open up new avenues in nuclear physics research which were presently hard to access. Such 'medium-resolution' spectroscopy has broad interest across several areas of contemporary interest such as the study of nuclear giant resonances. In addition to the connections to space science, it is striking that the demands of contemporary medical imaging have strong overlap with those of experimental nuclear physics. An example is the interest in PET-MRI combined imaging which requires putting scintillator detectors in a high magnetic field environment. This has led to strong advances in the area of silicon photomultipliers, a solid-state replacement for photomultiplier tubes, which are insensitive to magnetic fields. Broad application to nuclear physics of this technology may be foreseen. (paper)

  6. Absolute measurement of the responses of small lithium glass scintillators to gamma radiation

    International Nuclear Information System (INIS)

    Dalton, A.W.

    1987-04-01

    The absolute scintillation efficiency and intrinsic resolution of lithium glass scintillators for electron excitation have been determined over a range of electron energies, lithium concentrations and lithium enrichments. Measurements of these response characteristics form part of a study on the possible use of such glasses for the determination of tritium breeding in fusion reactor blanket experiments. The measurements were undertaken to establish a basis for extracting the information relating to tritium production reactions from the background signals induced within the glass scintillators by the neutron/gamma fields of a fusion reactor blanket. Criteria for the selection of glasses most suitable for tritium breeding measurements are discussed in tems of their observed responses

  7. Data process of liquid scintillation counting

    International Nuclear Information System (INIS)

    Ishikawa, Hiroaki; Kuwajima, Susumu.

    1975-01-01

    The use of liquid scintillation counting system has been significantly spread because automatic sample changers and printers have recently come to be incorporated. However, the system will be systematized completely if automatic data processing and the sample preparation of radioactive materials to be measured are realized. Dry or wet oxidation method is applied to the sample preparation when radioactive materials are hard to dissolve into scintillator solution. Since these several years, the automatic sample combustion system, in which the dry oxidation is automated, has been rapidly spread and serves greatly to labor saving. Since the printers generally indicate only counted number, data processing system has been developed, and speeded up calculating process, which automatically corrects quenching of samples for obtaining the final radioactivity required. The data processing system is roughly divided into on-line and off-line systems according to whether computers are connected directly or indirectly, while its hardware is classified to input, calculating and output devices. Also, the calculation to determine sample activity by external standard method is explained. (Wakatsuki, Y.)

  8. Silicon Photo-Multiplier Radiation Hardness Tests with a White Neutron Beam

    International Nuclear Information System (INIS)

    Montanari, A.; Tosi, N.; Pietropaolo, A.; Andreotti, M.; Baldini, W.; Calabrese, R.; Cibinetto, G.; Luppi, E.; Cotta Ramusino, A.; Malaguti, R.; Santoro, V.; Tellarini, G.; Tomassetti, L.; De Donato, C.; Reali, E.

    2013-06-01

    We report radiation hardness tests performed, with a white neutron beam, at the Geel Electron Linear Accelerator in Belgium on silicon Photo-Multipliers. These are semiconductor photon detectors made of a square matrix of Geiger-Mode Avalanche photo-diodes on a silicon substrate. Several samples from different manufacturers have been irradiated integrating up to about 6.2 x 10 9 1-MeV-equivalent neutrons per cm 2 . (authors)

  9. Scintillators

    International Nuclear Information System (INIS)

    Cusano, D.A.; Holub, F.F.; Prochazka, S.

    1979-01-01

    Scintillator bodies comprising phosphor materials and having high optical translucency with low light absorption, and methods of making the scintillator bodies, are described. Fabrication methods include (a) a hot-pressing process, (b) cold-pressing followed by sintering, (c) controlled cooling from a melt, and (d) hot-forging. The scintillator bodies that result are easily machined to desired shapes and sizes. Suitable phosphors include BaFCl:Eu, LaOBr:Tb, CsI:Tl, CaWO 4 and CdWO 4 . (U.K.)

  10. Prototype for a Radiation Hard Upgrade to the ATLAS ZDC

    CERN Document Server

    Phipps, Michael William; The ATLAS collaboration

    2017-01-01

    Increases in luminosity and collision energy at the LHC challenge the radiation hardness of detectors located along the beamline. This problem is especially acute for the Zero Degree Calorimeters (ZDCs) in ATLAS, which are exposed to around 10^{10} rad/yr, rendering the current version of the detector inviable during p+p running. To address this shortcoming and allow for important triggers and potential access to low-x physics, we designed a prototype detector that replaces quartz radiator material with a circulating, liquid hydrocarbon. It also features a dual-stage wavelength shifting scheme to transport light to silicon photo-multipliers, as well as both transverse and longitudinal segmentation to study the shower development in two dimensions. Design considerations, results from an SPS beam test and comparisons to GEANT simulation will be presented.

  11. Scintillation scanner

    International Nuclear Information System (INIS)

    Mehrbrodt, A.W.; Mog, W.F.; Brunnett, C.J.

    1977-01-01

    A scintillation scanner having a visual image producing means coupled through a lost motion connection to the boom which supports the scintillation detector is described. The lost motion connection is adjustable to compensate for such delays as may occur between sensing and recording scintillations. 13 claims, 5 figures

  12. Radiation detectors

    International Nuclear Information System (INIS)

    2013-01-01

    This sixth chapter presents the operational principles of the radiation detectors; detection using photographic emulsions; thermoluminescent detectors; gas detectors; scintillation detectors; liquid scintillation detectors; detectors using semiconductor materials; calibration of detectors; Bragg-Gray theory; measurement chain and uncertainties associated to measurements

  13. System tests of radiation hard optical links for the ATLAS semiconductor tracker

    International Nuclear Information System (INIS)

    Charlton, D.G.; Dowell, J.D.; Homer, R.J.; Jovanovic, P.; Kenyon, I.R.; Mahout, G.; Shaylor, H.R.; Wilson, J.A.; Rudge, A.; Fopma, J.; Mandic, I.; Nickerson, R.B.; Shield, P.; Wastie, R.; Weidberg, A.R.; Eek, L.-O.; Go, A.; Lund-Jensen, B.; Pearce, M.; Soederqvist, J.; Morrissey, M.; White, D.J.

    2000-01-01

    A prototype optical data and Timing, Trigger and Control transmission system based on LEDs and PIN-diodes has been constructed. The system would be suitable in terms of radiation hardness and radiation length for use in the ATLAS SemiConductor Tracker. Bit error rate measurements were performed for the data links and for the links distributing the Timing, Trigger and Control data from the counting room to the front-end modules. The effects of cross-talk between the emitters and receivers were investigated. The advantages of using Vertical Cavity Surface Emitting Lasers (VCSELs) instead of LEDs are discussed

  14. Comparison of two types of scintillation probe for moisture density gauge

    International Nuclear Information System (INIS)

    Machaj, B.

    1974-01-01

    Investigations of pulse shape discrimination scintillation probe, and amplitude discrimination probe as a detector for moisutre density gauge have been carried out. It was found that sandwich scintillator consisting of NE-421 + NE-102A was the best for pulse shape discrimination probe for thermal neutrons and gamma radiation detection. Similarly LiJ(Eu) crystal was the best for amplitude discrimination probe. The amplitude discrimination probe with LiJ(Eu) comparing to pulse shape discrimination probe with sandwich scintillator, provides approximately two times higher thermal neutron detection efficiency and higher count rate stability. It is considered therefore more suitable as the detector for moisture density gauge. (author)

  15. Luminescence and energy transfer in Garnet Scintillators

    NARCIS (Netherlands)

    Ogiegło, J.M.

    2012-01-01

    The thesis is focused on development and fundamental understanding of scintillators that play a central role in the field of medical imaging. These materials convert high energy, gamma or X-ray, radiation into visible light that is then used to create a detailed image of the patient’s body. The

  16. Evaluation of indigenously developed plastic scintillator sheet detector for surface radioactive contamination monitoring application

    International Nuclear Information System (INIS)

    Sahani, R.M.; Chaudhary, H.S.; Mahala, V.K.; Senwar, K.R.; Meena, J.P.

    2018-01-01

    Radioactive contamination may be caused by release of radioactivity in the environment due to accident at nuclear plant/reactor or spillage of loose radioactive materials in a laboratory. The protection of workers from potentially hazardous radiations emitted by the contaminants is a matter of prime concern. The detection of such radiations requires a monitoring system capable of measuring the level of radioactivity at the contaminated site. Plastic scintillators are widely used for large area radiation monitoring due to the ease of preparation in different shape and sizes. These detectors are sensitive to beta and gamma radiation therefore can be used for monitoring of beta and gamma contamination. In this paper, performance results of indigenously developed plastic scintillator sheet of area 800 cm 2 are reported

  17. Uncooled Radiation Hard SiC Schottky VUV Detectors Capable of Single Photon Sensing, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — This project seeks to design, fabricate, characterize and commercialize very large area, uncooled and radiative hard 4H-SiC VUV detectors capable of near single...

  18. Liquid scintillation solution

    International Nuclear Information System (INIS)

    Long, E.C.

    1977-01-01

    A liquid scintillation solution is described which includes (1) a scintillation solvent (toluene and xylene), (2) a primary scintillation solute (PPO and Butyl PBD), (3) a secondary scintillation solute (POPOP and Dimethyl POPOP), (4) a plurality of substantially different surfactants and (5) a filter dissolving and/or transparentizing agent. 8 claims

  19. Radiation hard silicon sensors for the CMS tracker upgrade

    CERN Document Server

    Pohlsen, Thomas

    2013-01-01

    At an instantaneous luminosity of $5 \\times 10^{34}$ cm$^{-2}$ s$^{-1}$, the high-luminosity phase of the Large Hadron Collider (HL-LHC) is expected to deliver a total of $3\\,000$ fb$^{-1}$ of collisions, hereby increasing the discovery potential of the LHC experiments significantly. However, the radiation dose of the tracking systems will be severe, requiring new radiation hard sensors for the CMS tracker. The CMS tracker collaboration has initiated a large material investigation and irradiation campaign to identify the silicon material and design that fulfils all requirements for detectors for the HL-LHC. Focussing on the upgrade of the outer tracker region, pad sensors as well as fully functional strip sensors have been implemented on silicon wafers with different material properties and thicknesses. The samples were irradiated with a mixture of neutrons and protons corresponding to fluences as expected for the positions of detector layers in the future tracker. Different proton energies were used for irr...

  20. Validation the quantification of beta emitters activity in urine by scintillation spectrometry in the liquid phase

    International Nuclear Information System (INIS)

    Sierra, I.; Hernandez, C.; Benito, P.; Lopez, C.

    2013-01-01

    In this paper the methodology used in the validation of the technique for quantifying activity of some beta emitters in urine ( 3 H, 1 4C, 3 5S, 3 2P and 9 0Sr) by scintillation spectrometry Liquid Phase (Liquid Scintillation Counting, LSC) is described in bio elimination Laboratory Service CIEMAT Radiation Dosimetry accredited since last year for carrying out assays measure radiation dose based on ISO forth above. (Author)

  1. Development of scintillation and luminescent detectors at BARC

    International Nuclear Information System (INIS)

    Pradhan, A.S.

    1991-01-01

    Research and development work carried out at the Bhabha Atomic Research Centre, Bombay, in the field of radiation detectors for various applications, particularly in the area of scintillation and luminescent detectors is reviewed. The review is presented in the form of 7 articles. (author). figs

  2. Development of radiophotometric dosemeters with high sensitivity using plastic scintillators as a light intensifier

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Hamada, M.M.

    1987-01-01

    Rectangular plates of plastic scintillators are developed and their effect as light converter evaluated, when used as film-holder in conventional photography dosemeters. In this dosemeter, the radiation that not interacts in the photographic film can be detected by light photons generation in the plastic scintillators, sensitizing the film. (C.G.C.) [pt

  3. Deep UV emitting scintillators for alpha and beta particle detection

    Energy Technology Data Exchange (ETDEWEB)

    Zhou, Y. [Department of Physics and Astronomy, University of Georgia, Athens, GA 30606 (United States); Jia, D.D.; Lewis, L.A. [Oak Ridge National Laboratory, Oak Ridge, TN 37831 (United States); Feofilov, S.P. [A.F. Ioffe Physical-Technical Institute, St. Petersburg, 194021 (Russian Federation); Meltzer, R.S., E-mail: rmeltzer@physast.uga.ed [Department of Physics and Astronomy, University of Georgia, Athens, GA 30606 (United States)

    2011-03-21

    Several deep UV emitting scintillators, whose emission falls in the solar blind region of the spectrum (200-280 nm), are described and their scintillator properties are characterized. They include LaPO{sub 4}:Pr, YPO{sub 4}:Pr, YAlO{sub 3}:Pr, Pr(PO{sub 3}){sub 3}, YPO{sub 4}:Bi and ScPO{sub 4}. These materials would facilitate the detection of ionizing radiation in open areas, even during the daylight hours, and could be used to support large area surveys that monitor for the presence of ionization radiation due, for example, to system leaks or transfer contamination. These materials can be used in the form of powders, thin films or paints for radiation detection. They are characterized for both beta radiation using electron beams (2-35 keV) and {sup 137}Cs and alpha radiations using {sup 241}Am sources. Their absolute light yields are estimated and are compared to that of Y{sub 2}SiO{sub 5}:Ce. Their light yields decrease as a function of electron energy but at 10 keV they approach 8000 ph/MeV.

  4. Scintillator Based Coded-Aperture Imaging for Neutron Detection

    International Nuclear Information System (INIS)

    Hayes, Sean-C.; Gamage, Kelum-A-A.

    2013-06-01

    In this paper we are going to assess the variations of neutron images using a series of Monte Carlo simulations. We are going to study neutron images of the same neutron source with different source locations, using a scintillator based coded-aperture system. The Monte Carlo simulations have been conducted making use of the EJ-426 neutron scintillator detector. This type of detector has a low sensitivity to gamma rays and is therefore of particular use in a system with a source that emits a mixed radiation field. From the use of different source locations, several neutron images have been produced, compared both qualitatively and quantitatively for each case. This allows conclusions to be drawn on how suited the scintillator based coded-aperture neutron imaging system is to detecting various neutron source locations. This type of neutron imaging system can be easily used to identify and locate nuclear materials precisely. (authors)

  5. Fiber-optic dosimeters for radiation therapy

    Science.gov (United States)

    Li, Enbang; Archer, James

    2017-10-01

    According to the figures provided by the World Health Organization, cancer is a leading cause of death worldwide, accounting for 8.8 million deaths in 2015. Radiation therapy, which uses x-rays to destroy or injure cancer cells, has become one of the most important modalities to treat the primary cancer or advanced cancer. The newly developed microbeam radiation therapy (MRT), which uses highly collimated, quasi-parallel arrays of x-ray microbeams (typically 50 μm wide and separated by 400 μm) produced by synchrotron sources, represents a new paradigm in radiotherapy and has shown great promise in pre-clinical studies on different animal models. Measurements of the absorbed dose distribution of microbeams are vitally important for clinical acceptance of MRT and for developing quality assurance systems for MRT, hence are a challenging and important task for radiation dosimetry. On the other hand, during the traditional LINAC based radiotherapy and breast cancer brachytherapy, skin dose measurements and treatment planning also require a high spatial resolution, tissue equivalent, on-line dosimeter that is both economical and highly reliable. Such a dosimeter currently does not exist and remains a challenge in the development of radiation dosimetry. High resolution, water equivalent, optical and passive x-ray dosimeters have been developed and constructed by using plastic scintillators and optical fibers. The dosimeters have peak edge-on spatial resolutions ranging from 50 to 500 microns in one dimension, with a 10 micron resolution dosimeter under development. The developed fiber-optic dosimeters have been test with both LINAC and synchrotron x-ray beams. This work demonstrates that water-equivalent and high spatial resolution radiation detection can be achieved with scintillators and optical fiber systems. Among other advantages, the developed fiber-optic probes are also passive, energy independent, and radiation hard.

  6. The influence of parotid gland sparing on radiation damages of dental hard tissues.

    Science.gov (United States)

    Hey, Jeremias; Seidel, Johannes; Schweyen, Ramona; Paelecke-Habermann, Yvonne; Vordermark, Dirk; Gernhardt, Christian; Kuhnt, Thomas

    2013-07-01

    The aim of the present study was to evaluate whether radiation damage on dental hard tissue depends on the mean irradiation dose the spared parotid gland is subjected to or on stimulated whole salivary flow rate. Between June 2002 and October 2008, 70 patients with neck and cancer curatively irradiated were included in this study. All patients underwent dental treatment referring to the guidelines and recommendations of the German Society of Dental, Oral and Craniomandibular Sciences prior, during, and after radiotherapy (RT). During the follow-up period of 24 months, damages on dental hard tissues were classified according to the RTOG/EORTC guidelines. The mean doses (D(mean)) during spared parotid gland RT were determined. Stimulated whole saliva secretion flow rates (SFR) were measured before RT and 1, 6, 12, 24 months after RT. Thirty patients showed no carious lesions (group A), 18 patients developed sporadic carious lesions (group B), and 22 patients developed general carious lesions (group C). Group A patients received a D mean of 21.2 ± 11.04 Gy. Group B patients received a D(mean) of 26.5 ± 11.59 Gy and group C patients received a D(mean) of 33.9 ± 9.93 Gy, respectively. The D(mean) of group A was significantly lower than the D(mean) of group C (p dental hard tissue correlates with increased mean irradiation doses as well as decreased salivary flow rates. Parotid gland sparing resulting in a dose below 20 Gy reduces radiation damage on dental hard tissues, and therefore, the dose may act as a predictor for the damage to be expected.

  7. Hardness variability in commercial technologies

    International Nuclear Information System (INIS)

    Shaneyfelt, M.R.; Winokur, P.S.; Meisenheimer, T.L.; Sexton, F.W.; Roeske, S.B.; Knoll, M.G.

    1994-01-01

    The radiation hardness of commercial Floating Gate 256K E 2 PROMs from a single diffusion lot was observed to vary between 5 to 25 krad(Si) when irradiated at a low dose rate of 64 mrad(Si)/s. Additional variations in E 2 PROM hardness were found to depend on bias condition and failure mode (i.e., inability to read or write the memory), as well as the foundry at which the part was manufactured. This variability is related to system requirements, and it is shown that hardness level and variability affect the allowable mode of operation for E 2 PROMs in space applications. The radiation hardness of commercial 1-Mbit CMOS SRAMs from Micron, Hitachi, and Sony irradiated at 147 rad(Si)/s was approximately 12, 13, and 19 krad(Si), respectively. These failure levels appear to be related to increases in leakage current during irradiation. Hardness of SRAMs from each manufacturer varied by less than 20%, but differences between manufacturers are significant. The Qualified Manufacturer's List approach to radiation hardness assurance is suggested as a way to reduce variability and to improve the hardness level of commercial technologies

  8. Design and test of a scintillation dosimeter for dosimetry measurements of high energy radiotherapy beams

    International Nuclear Information System (INIS)

    Fontbonne, J.M.

    2002-12-01

    This work describes the design and evaluation of the performances of a scintillation dosimeter developed for the dosimetry of radiation beams used in radiotherapy. The dosimeter consists in a small plastic scintillator producing light which is guided by means of a plastic optical fiber towards photodetectors. In addition to scintillation, high energy ionizing radiations produce Cerenkov light both in the scintillator and the optical fiber. Based on a wavelength analysis, we have developed a deconvolution technique to measure the scintillation light in the presence of Cerenkov light. We stress the advantages that are anticipated from plastic scintillator, in particular concerning tissue or water equivalence (mass stopping power, mass attenuation or mass energy absorption coefficients). We show that detectors based on this material have better characteristics than conventional dosimeters such as ionisation chambers or silicon detectors. The deconvolution technique is exposed, as well as the calibration procedure using an ionisation chamber. We have studied the uncertainty of our dosimeter. The electronics noise, the fiber transmission, the deconvolution technique and the calibration errors give an overall combined experimental uncertainty of about 0,5%. The absolute response of the dosimeter is studied by means of depth dose measurements. We show that absolute uncertainty with photons or electrons beams with energies ranging from 4 MeV to 25 MeV is less than ± 1 %. Last, at variance with other devices, our scintillation dosimeter does not need dose correction with depth. (author)

  9. Development of cryogenic Si detectors by CERN RD39 Collaboration for ultra radiation hardness in SLHC environment

    CERN Document Server

    Li, Z; Anbinderis, P; Anbinderis, T; D’Ambrosio, N; de Boer, Wim; Borchi, E; Borer, K; Bruzzi, M; Buontempo, S; Chen, W; Cindro, V; Dierlamm, A; Eremin, V; Gaubas, E; Gorbatenko, V; Grigoriev, E; Hauler, F; Heijne, Erik H M; Heising, S; Hempel, O; Herzog, R; Härkönen, J; Ilyashenko, I; Janos, S; Jungermann, L; Kalesinskas, V; Kapturauskas, J; Laiho, R; Luukka, P; Mandic, I; De Masi, R; Menichelli, D; Mikuz, M; Militaru, O; Niinikosky, T O; O’Shea, V; Pagano, S; Paul, S; Piotrzkowski, K; Pretzl, K; Rato-Mendes, P; Rouby, X; Ruggiero, G; Smith, K; Sonderegger, P; Sousa, P; Tuominen, E; Tuovinen, E; Verbitskaya, E; Vaitkus, J; Wobst, E; Zavrtanik, M

    2007-01-01

    There are two key approaches in our CERN RD 39 Collaboration efforts to obtain ultra-radiation-hard Si detectors: (1) use of the charge/current injection to manipulate the detector internal electric field in such a way that it can be depleted at a modest bias voltage at cryogenic temperature range (150 K), and (2) freezing out of the trapping centers that affects the CCE at cryogenic temperatures lower than that of the liquid nitrogen (LN2) temperature. In our first approach, we have developed the advanced radiation hard detectors using charge or current injection, the current injected diodes (CID). In a CID, the electric field is controlled by injected current, which is limited by the space charge, yielding a nearly uniform electric field in the detector, independent of the radiation fluence. In our second approach, we have developed models of radiation-induced trapping levels and the physics of their freezing out at cryogenic temperatures.

  10. Photonic Crystals: Enhancing the Light Output of Scintillation Based Detectors

    CERN Document Server

    Knapitsch, Arno Richard

    A scintillator is a material which emits light when excited by ionizing radiation. Such materials are used in a diverse range of applications; From high energy particle physics experiments, X-ray security, to nuclear cameras or positron emission tomography. Future high-energy physics (HEP) experiments as well as next generation medical imaging applications are more and more pushing towards better scintillation characteristics. One of the problems in heavy scintillating materials is related to their high index of refraction. As a consequence, most of the scintillation light produced in the bulk material is trapped inside the crystal due to total internal reflection. The same problem also occurs with light emitting diodes (LEDs) and has for a long time been considered as a limiting factor for their overall efficiency. Recent developments in the area of nanophotonics were showing now that those limitations can be overcome by introducing a photonic crystal (PhC) slab at the outcoupling surface of the substrate. P...

  11. Radiation-Hard Complementary Integrated Circuits Based on Semiconducting Single-Walled Carbon Nanotubes.

    Science.gov (United States)

    McMorrow, Julian J; Cress, Cory D; Gaviria Rojas, William A; Geier, Michael L; Marks, Tobin J; Hersam, Mark C

    2017-03-28

    Increasingly complex demonstrations of integrated circuit elements based on semiconducting single-walled carbon nanotubes (SWCNTs) mark the maturation of this technology for use in next-generation electronics. In particular, organic materials have recently been leveraged as dopant and encapsulation layers to enable stable SWCNT-based rail-to-rail, low-power complementary metal-oxide-semiconductor (CMOS) logic circuits. To explore the limits of this technology in extreme environments, here we study total ionizing dose (TID) effects in enhancement-mode SWCNT-CMOS inverters that employ organic doping and encapsulation layers. Details of the evolution of the device transport properties are revealed by in situ and in operando measurements, identifying n-type transistors as the more TID-sensitive component of the CMOS system with over an order of magnitude larger degradation of the static power dissipation. To further improve device stability, radiation-hardening approaches are explored, resulting in the observation that SWNCT-CMOS circuits are TID-hard under dynamic bias operation. Overall, this work reveals conditions under which SWCNTs can be employed for radiation-hard integrated circuits, thus presenting significant potential for next-generation satellite and space applications.

  12. Radiation hardness tests of SiPMs for the JLab Hall D Barrel calorimeter

    International Nuclear Information System (INIS)

    Qiang, Yi; Zorn, Carl; Barbosa, Fernando; Smith, Elton

    2013-01-01

    We report on the measurement of the neutron radiation hardness of silicon photomultipliers (SiPMs) manufactured by Hamamatsu Corporation in Japan and SensL in Ireland. Samples from both companies were irradiated by neutrons created by a 1 GeV electron beam hitting a thin lead target at Jefferson Lab Hall A. More tests regarding the temperature dependence of the neutron radiation damage and self-annealing were performed on Hamamatsu SiPMs using a calibrated Am–Be neutron source from the Jefferson Lab Radiation Control group. As the result of irradiation both dark current and dark rate increase linearly as a function of the 1 MeV equivalent neutron fluence and a temperature dependent self-annealing effect is observed.

  13. Numerical Simulation of Radial and Angular Distribution of γ-Ray's Energy Deposition in Scintillation Optical Fibre

    International Nuclear Information System (INIS)

    Tang Shibiao; Yin Zejie; Tang Yu; Huang Huan

    2006-01-01

    Angular and radial distributions of the energy deposition of γ-ray radiation in scintillation optical fibres are simulated and analysed using the Geant4 system. The results show a linear relation between the energy deposition and the radius of the fibres. The deposition is roughly inversely proportional to sinθ with θ the incident angle relative to the fibre axis. The results could provide corrections to the measurements of the scintillation fibres used in monitoring the γ-ray radiation

  14. Radiation hardness of WLS fibres for the ATLAS Tile Calorimeter

    CERN Document Server

    David, M; Maio, A

    2007-01-01

    In this document we present the data obtained in the irradiation in a Co-60 source of WLS fibers for the TileCal calorimeter. The optical, mechanical and radiation hardness properties of these fibers were developed in close contact with three producers: Bicron, Kuraray and Pol.Hi.Tech. The results on the degradation of the light output and attenuation length from five irradiations are presented. The fibers were irradiated with a total dose at least 3 times higher than the dose predicted for 10 years of operation of LHC at nominal luminosity.

  15. Liquid scintillation solution

    International Nuclear Information System (INIS)

    Long, E.C.

    1976-01-01

    The invention deals with a liquid scintillation solution which contains 1) a scintillation solvent (toluol), 2) a primary scintillation solute (PPO), 3) a secondary scintillation solute (dimethyl POPOP), 4) several surfactants (iso-octyl-phenol polyethoxy-ethanol and sodium di-hexyl sulfosuccinate) essentially different from one another and 5) a filter resolution and/or transparent-making agent (cyclic ether, especially tetrahydrofuran). (HP) [de

  16. Temperature dependence of scintillation properties of SrMoO{sub 4}

    Energy Technology Data Exchange (ETDEWEB)

    Mikhailik, V.B., E-mail: vmikhai@hotmail.com [Diamond Light Source, Harwell Science Campus, Didcot OX11 0DE (United Kingdom); Elyashevskyi, Yu. [Department of Physics, University of Oxford, Keble Rd., Oxford OX1 3RH (United Kingdom); Scientific-technical and Educational Centre of Low Temperature Studies, I. Franko National University of Lviv, 50 Dragomanova Str., 79005 Lviv (Ukraine); Kraus, H. [Department of Physics, University of Oxford, Keble Rd., Oxford OX1 3RH (United Kingdom); Kim, H.J. [Department of Physics of Kyungpook National University, 1370 Sangyeok-dong, Buk-gu, Daegu 702-701 (Korea, Republic of); Kapustianyk, V.; Panasyuk, M. [Scientific-technical and Educational Centre of Low Temperature Studies, I. Franko National University of Lviv, 50 Dragomanova Str., 79005 Lviv (Ukraine)

    2015-08-21

    Studies of the X-ray luminescence and scintillation properties of a SrMoO{sub 4} crystal as function of temperature down to T=10 K have been carried out. The luminescence in SrMoO{sub 4} is quenched at room temperature, but below T<200 K the crystal exhibits a broad emission band with a maximum at a wavelength of 520 nm. The emission is attributed to the radiative decay of self-trapped excitons and defects acting as traps for the exactions at low temperatures. Such complex character of radiative decay is reflected in the kinetics which contains several components plus a contribution from delayed recombination at low temperatures. The temperature dependence of scintillation light output of SrMoO{sub 4} was studied. Comparing with a reference ZnWO{sub 4} crystal measured under the same experimental conditions it was found that the light output of SrMoO{sub 4} is 15±5%. It is suggested, therefore, that there is scope for optimisation of strontium molybdate for application as scintillator in cryogenic rare event searches.

  17. Investigation of linear accelerator pulse delivery using fast organic scintillator measurements

    DEFF Research Database (Denmark)

    Beierholm, Anders Ravnsborg; Andersen, Claus Erik; Lindvold, Lars René

    2010-01-01

    Fiber-coupled organic plastic scintillators present an attractive method for time-resolved dose measurements during radiotherapy. Most organic scintillators exhibit a fast response, making it possible to use them to measure individual high-energy X-ray pulses from a medical linear accelerator...... performed on Varian medical linear accelerators, delivering 6 MV X-ray beams. The dose delivery per radiation pulse was found to agree with expectations within roughly 1%, although minor discrepancies and transients were evident in the measurements....

  18. Studies of radiation hardness of MOS devices for application in a linear collider vertex detector

    Energy Technology Data Exchange (ETDEWEB)

    Wei, Qingyu

    2008-10-17

    The proposed International Linear Collider (ILC) together with the Large Hadron Collider (LHC) at CERN serve as a combined tool to explore the mysteries of the universe: the former is a precision machine and the latter can be considered as a finding machine. The key component of the ILC is the vertex detector that should be placed as close as possible to the Interaction Point (IP) and has better radiation tolerance against the dominant electron-positron pair production background from beam-beam interactions. A new generation of MOS-type Depleted-Field-Effect Transistor (MOSDEPFET) active pixel detectors has been proposed and developed by Semiconductor Labor Munich for Physics and for extraterrestrial Physics in order to meet the requirements of the vertex detector at the ILC. Since all MOS devices are susceptible to ionizing radiation, the main topic is focused on the radiation hardness of detectors, by which a series of physical processes are analyzed: e.g. surface damage due to ionizing radiation as well as damage mechanisms and their associated radiation effects. As a consequence, the main part of this thesis consists of a large number of irradiation experiments and the corresponding discussions. Finally, radiation hardness of the detectors should be improved through a set of concluded experiences that are based on a series of analysis of the characteristic parameters using different measurement techniques. The feasibility of the MOSDEPFET-based vertex detector is, therefore, predicted at ILC. (orig.)

  19. Studies of radiation hardness of MOS devices for application in a linear collider vertex detector

    International Nuclear Information System (INIS)

    Wei, Qingyu

    2008-01-01

    The proposed International Linear Collider (ILC) together with the Large Hadron Collider (LHC) at CERN serve as a combined tool to explore the mysteries of the universe: the former is a precision machine and the latter can be considered as a finding machine. The key component of the ILC is the vertex detector that should be placed as close as possible to the Interaction Point (IP) and has better radiation tolerance against the dominant electron-positron pair production background from beam-beam interactions. A new generation of MOS-type Depleted-Field-Effect Transistor (MOSDEPFET) active pixel detectors has been proposed and developed by Semiconductor Labor Munich for Physics and for extraterrestrial Physics in order to meet the requirements of the vertex detector at the ILC. Since all MOS devices are susceptible to ionizing radiation, the main topic is focused on the radiation hardness of detectors, by which a series of physical processes are analyzed: e.g. surface damage due to ionizing radiation as well as damage mechanisms and their associated radiation effects. As a consequence, the main part of this thesis consists of a large number of irradiation experiments and the corresponding discussions. Finally, radiation hardness of the detectors should be improved through a set of concluded experiences that are based on a series of analysis of the characteristic parameters using different measurement techniques. The feasibility of the MOSDEPFET-based vertex detector is, therefore, predicted at ILC. (orig.)

  20. Scintillation screen materials for beam profile measurements of high energy ion beams

    Energy Technology Data Exchange (ETDEWEB)

    Krishnakumar, Renuka

    2016-06-22

    For the application as a transverse ion beam diagnostics device, various scintillation screen materials were analysed. The properties of the materials such as light output, image reproduction and radiation stability were investigated with the ion beams extracted from heavy ion synchrotron SIS-18. The ion species (C, Ne, Ar, Ta and U) were chosen to cover the large range of elements in the periodic table. The ions were accelerated to the kinetic energies of 200 MeV/u and 300 MeV/u extracted with 300 ms pulse duration and applied to the screens. The particle intensity of the ion beam was varied from 10{sup 4} to 10{sup 9} particles per pulse. The screens were irradiated with typically 40 beam pulses and the scintillation light was captured using a CCD camera followed by characterization of the beam spot. The radiation hardness of the screens was estimated with high intensity Uranium ion irradiation. In the study, a linear light output for 5 orders of magnitude of particle intensities was observed from sensitive scintillators and ceramic screens such as Al{sub 2}O{sub 3}:Cr and Al{sub 2}O{sub 3}. The highest light output was recorded by CsI:Tl and the lowest one by Herasil. At higher beam intensity saturation of light output was noticed from Y and Mg doped ZrO{sub 2} screens. The light output from the screen depends not only on the particle intensity but also on the ion species used for irradiation. The light yield (i.e. the light intensity normalised to the energy deposition in the material by the ion) is calculated from the experimental data for each ion beam setting. It is shown that the light yield for light ions is about a factor 2 larger than the one of heavy ions. The image widths recorded exhibit a dependence on the screens material and differences up to 50 % were registered. On radiation stability analysis with high particle intensity of Uranium ions of about 6 x 10{sup 8} ppp, a stable performance in light output and image reproduction was documented from Al

  1. Calibration of LiBaF sub 3 Ce scintillator for fission spectrum neutrons

    CERN Document Server

    Reeder, P L

    2002-01-01

    The scintillator LiBaF sub 3 doped with small amounts of Ce sup + sup 3 has the ability to distinguish heavy charged particles (p, d, t, or alpha) from beta and/or gamma radiation based on the presence or absence of nanosecond components in the scintillation light output. Since the neutron capture reaction on sup 6 Li produces recoil alphas and tritons, this scintillator also discriminates between neutron induced events and beta or gamma interactions. An experimental technique using a time-tagged sup 2 sup 5 sup 2 Cf source has been used to measure the efficiency of this scintillator for neutron capture, the calibration of neutron capture pulse height, and the pulse height resolution--all as a function of incident neutron energy.

  2. Method for determining efficiency in a liquid scintillation system

    International Nuclear Information System (INIS)

    Laney, B.H.

    1975-01-01

    This invention relates to a method of counting radioactive events in a liquid scintillation radiation detecting and counting apparatus by utilizing pulses generated by a photomultiplying means resulting from scintillations caused by radioactive events. A counting efficiency value is assigned to each pulse generated in the photomultiplying means according to the height of the pulse. The numerical inverse of each assigned counting efficiency value is determined and each numerical inverse is recorded as an actual number of radioactive events with each having a pulse height identical to that of the corresponding pulse generated in the photomultiplying means. (Patent Office Record)

  3. Isotope weigher with scintillation detector in transmission geometry

    International Nuclear Information System (INIS)

    Machaj, B.; Antoniak, W.

    1976-01-01

    Measuring head of isotope conveyor belt weigher contains gamma radiation source 137 Cs or 60 Co having activity less than 5 mCi, long plastic scintillator as a radiation detector and tachometer as a belt speed sensor. Measurement of instantaneous belt loading is based on the principle of radiation attenuation in transmission geometry. Electronic signal originated by gamma radiation is first linearized and then multiplied by signal which is proportional to belt speed, thus giving at the multiplier output signal which is proportional to the mass being conveyed. Mass weighing accuracy is not worse than 2%. Belt width of conveyor belt 30 - 180 cm. (author)

  4. Radon gamma-ray spectrometry with YAP:Ce scintillator

    CERN Document Server

    Plastino, W; De Notaristefani, F

    2002-01-01

    The detection properties of a YAP:Ce scintillator (YAlO sub 3 :Ce crystal) optically coupled to a Hamamatsu H5784 photomultiplier with standard bialkali photocathode have been analyzed. In particular, the application to radon and radon-daughters gamma-ray spectrometry was investigated. The crystal response has been studied under severe extreme conditions to simulate environments of geophysical interest, particularly those found in geothermal and volcanic areas. Tests in water up to a temperature of 100 deg.C and in acids solutions such as HCl (37%), H sub 2 SO sub 4 (48%) and HNO sub 3 (65%) have been performed. The measurements with standard radon sources provided by the National Institute for Metrology of Ionizing Radiations (ENEA) have emphasized the non-hygroscopic properties of the scintillator and a small dependence of the light yield on temperature and HNO sub 3. The data collected in this first step of our research have pointed out that the YAP:Ce scintillator can allow high response stability for rad...

  5. Development of radiation monitoring and visualization systems for Fukushima. GPS monitoring system, Dose3DMap system, and LED-coupled scintillating fiber detector

    International Nuclear Information System (INIS)

    Nakao, Noriaki; Kosako, Kazuaki; Kinoshita, Norikazu; Kawaguchi, Masato

    2016-01-01

    Lands that were contaminated with radioactive elements following the Fukushima Daiichi Nuclear Power Plant accident in 2011 have been decontaminated, and the construction of an interim storage facility for radioactive waste is planned. A GPS monitoring system was developed to concomitantly determine a location and measure the radiation level at the location. Moreover, a mapping system that produces radiation maps at the measurement locations and also predicts post-decontamination radiation maps using the compiled Monte Carlo simulation program was constructed. These systems were used for decontamination planning and estimation of the decontamination effect. An LED-coupled scintillating fiber detector was developed for visually monitoring radiation in real time at the interim storage facility. The LEDs display different colors corresponding to different radiation levels at the measurement locations along the fiber detector, the maximum length of which is 50 m. Thus, the radiation levels at all positions along the length of the detector can be visually monitored in real time. Moreover, it is useful for radiation safety and for risk communication with radiation workers and residents close to the site. (author)

  6. Study on fast luminescence component induced by gamma-rays in Ce doped LiCaAlF6 scintillators

    International Nuclear Information System (INIS)

    Watanabe, Kenichi; Kondo, Yoshiyuki; Yamazaki, Atsushi; Uritani, Akira; Iguchi, Tetsuo; Kawaguchi, Noriaki; Fukuda, Kentaro; Ishizu, Sumito; Yanagida, Takayuki; Fujimoto, Yutaka; Yoshikawa, Akira

    2014-01-01

    We discuss the origin of the fast luminescence component induced by fast electrons generated in gamma-ray interactions in Ce doped LiCaAlF 6 scintillators. Although the slow luminescence component induced by Ce 3+ emissions depends on the Ce concentration in the LiCaAlF 6 scintillator, the fast component is independent. The fast component is suggested to be generated in the host matrix of the LiCaAlF 6 crystal. From quantitative considerations based on Frank–Tamm equation, which shows the light yield of the Cherenkov radiation, the Cherenkov radiation was determined as the origin of the fast component. We, additionally, found that the slow rise time of main Ce 3+ emissions in the Ce:LiCaAlF 6 scintillator plays an important role to perform the pulse shape discrimination. - Highlights: • The fast luminescence in Ce:LiCaAlF 6 scintillator is generated in the host matrix. • The origin of the fast luminescence is determined as the Cherenkov radiation. • The slow rise time also plays an important role to perform PSD

  7. Hydrogenated amorphous silicon radiation detectors: Material parameters; radiation hardness; charge collection

    International Nuclear Information System (INIS)

    Qureshi, S.

    1991-01-01

    Properties of hydrogenated amorphous silicon p-i-n diodes relevant to radiation detection applications were studied. The interest in using this material for radiation detection applications in physics and medicine was motivated by its high radiation hardness and the fact that it can be deposited over large area at relatively low cost. Thick, fully depleted a-Si:H diodes are required for sufficient energy deposition by a charged particle and better signal to noise ratio. A sizeable electric field is essential for charge collection in a -Si:H diodes. The large density of ionized defects that exist in the i layer when the diode is under DC bias causes the electric field to be uniform. Material parameters, namely carrier mobility and lifetime and the ionized defect density in thick a-Si:H p-i-n diodes were studied by the transient photoconductivity method. The increase in diode leakage current with reverse bias over the operating bias was consistent with the Poole-Frenkel effect, involving excitation of carriers from neutral defects. The diode noise over the operating voltage range was completely explained in terms of the shot noise component for CR-(RC) 4 (pseudo-Gaussian) shaping at 3 μs shaping time and the noise component at 0 V bias (delta and thermal noise) added in quadrature. Irradiation with 1 Mev neutrons produced no significant degradation in leakage current and noise at fluences exceeding 4 x 10 14 cm -2 . Irradiation with 1.4 Mev proton fluence of 1 x 10 14 cm -2 decreased carrier lifetime by a factor of ∼4. Degradation in leakage current and noise became significant at proton fluence of ∼10 13 cm -2

  8. Radiation distribution sensor with optical fibers for high radiation fields

    International Nuclear Information System (INIS)

    Takada, Eiji; Kimura, Atsushi; Hosono, Yoneichi; Takahashi, Hiroyuki; Nakazawa, Masaharu

    1999-01-01

    Radiation distribution sensors with their feasibilities have been described in earlier works. However, due to large radiation induced transmission losses in optical fibers, especially in the visible wavelength region, it has been difficult to apply these techniques to high radiation fields. In this study, we proposed a new concept of optical fiber based radiation distribution measurements with near infrared (IR) emission. Near IR scintillators were attached to the ends of optical fibers, where the fibers were bundled and connected to an N-MOS line sensor or a cooled CCD camera. From the measurements of each area density, the radiation levels at the positions of the scintillators can be known. The linearity between the gamma dose rate at each scintillator and the registered counts has been examined. For correcting the radiation induced loss effects, we applied the Optical Time Domain Reflectometry technique to measure the loss distribution and from the results, a possibility for correction of the loss effect has been demonstrated. The applicable dose rate range was evaluated to be from 0.1 to 10 3 Gy/h. This system can be a promising tool as a flexible dose rate distribution monitor in radiation facilities like nuclear plants and accelerator facilities. (author)

  9. Reconstruction of scintillations coordinates in cylindrical large detector with radial arrangement of photomultipliers

    CERN Document Server

    Alekseev, V M; Verbitskij, V S; Verbitskij, S S; Lapik, M A; Tselebrovskij, A N; Lapik, A M; Rusakov, A V; Savopulo, M L; Smirnov, V V; Chubarov, M N

    2002-01-01

    Paper describes algorithm to calculate scintillation coordinates elaborated for a coordinate-sensitive large detector based on liquid scintillator with radial arrangement of nine FEU-174 photomultipliers. Paper contains the simulation results of coordinate resolution dependence in detector centre on energy of gamma-quanta within 0.2-8 MeV range. Using this algorithm one processed the experimental data obtained with application of sup 6 sup 0 Co gamma-radiation source and the results of determination of scintillation coordinates for gamma-quanta beam. The accuracy of coordinate determination in detector centre when applying the descried algorithm constitutes approx 10 mm

  10. Development of plastic scintillators for use in the field of radioprotection and environmental monitoring

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Hamada, M.M.; Chandra, U.

    1986-01-01

    A plastic scintillator detector based on available know how, for applications in radiological protection - whole body counting, continuous monitoring of air and measurement of low level alpha and beta radioactivity environmental samples is described. Results of method of preparation of the plastic scintillator using styrene monomer and PPO (scintillator), POPOP (wave lenght shifter) are described. For measurement of gamma radiation point sources of 60 Co, 137 Cs and 125 I were used and detected with plastic scintillator detectors prepared with various concentrations (% by weight) of PPO. The data of relative pulse heights are presented. For measurement of alpha and beta radioactivity in environmental samples, thin discs of plastic detector were machined and polished. (Author) [pt

  11. Development of a Compact Gamma-ray Detector for a Neural-Network Radiation Monitoring

    Energy Technology Data Exchange (ETDEWEB)

    Kim, H. S.; Ha, J. H.; Lee, K. H. [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of); Lee, C. H. [Hanyang Univ., Seoul (Korea, Republic of)

    2012-03-15

    Radiation monitoring is very important to secure safety in nuclear-related facilities and against nuclear terrorism. For wide range of radiation monitoring, neutral network system of radiation detection is most efficient way. Thus, a compact radiation detector is useful to install in wide range to be concerned. A compact gamma-ray detector was fabricated by using a CsI(Tl) scintillator, which was matched with the formerly developed PIN photodiode, for a neural network radiation monitoring. At room temperature, the fabricated compact gamma-ray detector demonstrates an energy resolution of 13.3 % for 662 keV 6.9% for 1330 keV. The compactness, the low-voltage power consumption and the physical hardness are very useful features for a neural network radiation monitoring. In this study, characteristics of a fabricated compact gamma-ray detector were presented. An important aspect to consider in a neural-network radiation monitoring such as reaction probability of the fabricated compact detector for angle of incident gamma-ray was also addressed.

  12. Investigation on n/γ discrimination methods for liquid scintillator detector

    International Nuclear Information System (INIS)

    Li Kuinian; Li Yang; Zhang Mei; Zhang Zhongbing; Li Binkang; Zhang Xiaodong; Liu Jun; Zhang Xianpeng

    2014-01-01

    To obtain the n/γ discrimination ability of different digital pulse shape discrimination methods, four methods (rising time method, charge comparison method, pulse gradient analysis and frequency gradient analysis) in americium-beryllium mixed radiation fields were demonstrated. The signals from EJ-301 and BC501A scintillator detectors were digitized using oscilloscope. A comparison was taken among the four discrimination methods. The discrimination results of the four methods in liquid scintillator detectors show that the rising time method is the best and it provides a good choice in real-time n/γ discrimination system. (authors)

  13. Calibration of the neutron scintillation counter threshold

    International Nuclear Information System (INIS)

    Noga, V.I.; Ranyuk, Yu.N.; Telegin, Yu.N.

    1978-01-01

    A method for calibrating the threshold of a neutron counter in the form of a 10x10x40 cm plastic scintillator is described. The method is based on the evaluation of the Compton boundary of γ-spectrum from the discrimination curve of counter loading. The results of calibration using 60 Co and 24 Na γ-sources are given. In order to eValuate the Compton edge rapidly, linear extrapolation of the linear part of the discrimination curve towards its intersection with the X axis is recommended. Special measurements have shown that the calibration results do not practically depend on the distance between the cathode of a photomultiplier and the place where collimated γ-radiation of the calibration source reaches the scintillator

  14. Compound Semiconductor Radiation Detector

    International Nuclear Information System (INIS)

    Kim, Y. K.; Park, S. H.; Lee, W. G.; Ha, J. H.

    2005-01-01

    In 1945, Van Heerden measured α, β and γ radiations with the cooled AgCl crystal. It was the first radiation measurement using the compound semiconductor detector. Since then the compound semiconductor has been extensively studied as radiation detector. Generally the radiation detector can be divided into the gas detector, the scintillator and the semiconductor detector. The semiconductor detector has good points comparing to other radiation detectors. Since the density of the semiconductor detector is higher than that of the gas detector, the semiconductor detector can be made with the compact size to measure the high energy radiation. In the scintillator, the radiation is measured with the two-step process. That is, the radiation is converted into the photons, which are changed into electrons by a photo-detector, inside the scintillator. However in the semiconductor radiation detector, the radiation is measured only with the one-step process. The electron-hole pairs are generated from the radiation interaction inside the semiconductor detector, and these electrons and charged ions are directly collected to get the signal. The energy resolution of the semiconductor detector is generally better than that of the scintillator. At present, the commonly used semiconductors as the radiation detector are Si and Ge. However, these semiconductor detectors have weak points. That is, one needs thick material to measure the high energy radiation because of the relatively low atomic number of the composite material. In Ge case, the dark current of the detector is large at room temperature because of the small band-gap energy. Recently the compound semiconductor detectors have been extensively studied to overcome these problems. In this paper, we will briefly summarize the recent research topics about the compound semiconductor detector. We will introduce the research activities of our group, too

  15. Scintillation counter: photomultiplier tube alignment

    International Nuclear Information System (INIS)

    Olson, R.E.

    1975-01-01

    A scintillation counter, particularly for counting gamma ray photons, includes a massive lead radiation shield surrounding a sample-receiving zone. The shield is disassembleable into a plurality of segments to allow facile installation and removal of a photomultiplier tube assembly, the segments being so constructed as to prevent straight-line access of external radiation through the shield into the sample receiving zone. Provisions are made for accurately aligning the photomultiplier tube with respect to one or more sample-transmitting bores extending through the shield to the sample receiving zone. A sample elevator, used in transporting samples into the zone, is designed to provide a maximum gamma-receiving aspect to maximize the gamma detecting efficiency. (auth)

  16. Apllication of plastic scintillators with light intensifiers in photographic dosemeters to increase the sensitivity

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Hamada, M.M.

    1988-01-01

    The purpose of this work is to evaluate the effect of a light converter (plastic scintillator) to use as badge in the conventional photographic dosimeter. In this dosimeter the light is a new component of emulsion sensibilization. So, even radiations that are not able to reach the photographic emulsion can be detected by producing photons in the plastic scintillator. (author) [pt

  17. Light collection and its fluctuation in Cherenkov and scintillation spectrometers

    International Nuclear Information System (INIS)

    Kitaev, D.F.; Samedov, V.V.; Stolyarova, E.L.

    1982-01-01

    The GAMMA program for calculating light collection in the Cherenkov and scintillation counters is described. Together with the shower modelling program the GAMNA program can be used for evaluating the output signal and energy resolution of shower spectrometers. Principle formulae and block diagram of the program are given. Results of test calculations performed on the example of scintillation counters of culindrical and rectangular shapes were considered. Modelling of the radiation polarization envisaged in the program permits to take account of the effect of selective discrimination of photoelectron amplifier photocathode. The program analyzes, for the present situation, calculation errors which permits to plan in advance the calculation with the given accuracy. The program permits to use additional subprograms together with it where it is possible to take account of other peculiarities of light collection, for example, the presence of outer reflectors and focusing elements of light collection systems, particle slowing down in the spectrometer radiator expressed in the change of angle of semiaperture of the Cherenkov radiation cone. It is concluded on the basis of analyzing results of test calculations that the choosen technique and algorithms of light collection coefficient calculation in spectrometer radiators are correct

  18. Development of self-propelled measuring system for 2-dimensional distribution of radiation beam using plastic scintillation fibers

    International Nuclear Information System (INIS)

    Matsumura, Shuji; Kitahara, Sigeo; Yamanishi, Akio; Nose, Hiroyuki; Tisaka, Osamu

    2013-01-01

    Conventional 2-dimensional distribution of radiation beam is usually estimated from dose rates on a lot of dispersed spots, which has two problems. One is that it takes much time to measure distribution in a large area, and another problem is it is difficult to detect a localized hot spot from dispersed measurement results. To solve these problems we have developed a self-propelled measuring system adopting plastic scintillation fibers (PSF) as a detector. Estimating dose distribution in PSF and scanning PSF with self-propelled system give a 2-dimensional distribution of radiation beam in shorter measuring time and better spatial resolution than usual. A global positioning system was also installed to our system to know the absolute position of interest. With this system we have verified that we can estimate the 2-dimensional distribution in area of 2,000 m 2 in an hour. This report describes the overview of our newly developed system. (author)

  19. Scintillator performance at low dose rates and low temperatures for the CMS High Granularity Calorimeter for HL-LHC

    CERN Document Server

    Ricci-Tam, Francesca

    2018-01-01

    The High Luminosity LHC (HL-LHC) will integrate 10 times more luminosity than the LHC, posing significant challenges for radiation tolerance, especially for forward calorimetry, and highlights the issue for future colliders. As part of its HL-LHC upgrade program, the CMS collaboration is designing a High Granularity Calorimeter to replace the existing endcap calorimeters. The upgrade includes both electromagnetic and hadronic components, with the latter using a mixture of silicon sensors (in the highest radiation regions at high pseudorapidity) and scintillator as its active components. The scintillator will nevertheless receive large doses accumulated at low dose rates, and will have to operate at low temperature - around -30 degrees Celsius. We discuss measurements of scintillator radiation tolerance, from in-situ measurements from the current CMS endcap calorimeters, and from measurements at low temperature and low dose-rate at gamma sources in the laboratory.

  20. SU-F-T-239: Evaluation of Plastic Scintillator Light Output for Various Lead Doping Concentrations: Towards LET Detection

    International Nuclear Information System (INIS)

    Nusrat, H; Pang, G; Sarfehnia, A

    2016-01-01

    Purpose: This work seeks to develop a beam quality meter using multiple differently doped plastic scintillators that are thus intrinsically beam-quality dependent. Plastic scintillators spontaneously emit visible light upon irradiation; the amount of light produced is dependent on stopping power (closely related to LET) according to Birks’ law. Doping plastic scintillators can be used to tune their sensitivity to specific LET ranges. Methods: GEANT4.10.1 Monte Carlo (MC) was used to evaluate the response of various scintillator dopant combinations. MC radiation transport and scintillator light response were validated against previously published literature. Current work involves evaluating detector response experimentally; to that end, a detector prototype with interchangeable scintillator housing was constructed. Measurement set-up guides light emitted by the scintillator to a photomultiplier tube via a glass taper junction coupled to an optical fiber. The resulting signal is measured by an electrometer, and normalized to dose readout from a diode. Measurements have been done using clinical electron and orthovoltage beams. MC response (simulated scintillator light normalized to dose scored inside the scintillating volume) was evaluated for four different LET radiations for an undoped and 1%Pb doped scintillator (σ=0.85%). Simulated incident electrons included: 0.05, 0.1, 0.2, 6, 12, and 18 MeV; these energies correspond to a range of stopping power (related to LET) values ranging from 1.824 to 11.09 MeVcm"2g"−"1 (SCOL from NIST-ESTAR). Results: Initial MC results show a distinct divergence in scintillator response as LET increases. The response for undoped plastic scintillator indicated a 35.0% increase in signal when going from 18 MeV (low LET) to 0.05 MeV (high LET) while 1%-Pb doped scintillator indicated a 100.9% increase. Conclusion: After validating MC against measurement, simulations will be used to test various concentrations (2%, 4%, 6%) of different

  1. SU-F-T-239: Evaluation of Plastic Scintillator Light Output for Various Lead Doping Concentrations: Towards LET Detection

    Energy Technology Data Exchange (ETDEWEB)

    Nusrat, H [Ryerson University, Toronto, ON (Canada); Pang, G; Sarfehnia, A [Sunnybrook Health Sciences Centre, Toronto, ON (Canada)

    2016-06-15

    Purpose: This work seeks to develop a beam quality meter using multiple differently doped plastic scintillators that are thus intrinsically beam-quality dependent. Plastic scintillators spontaneously emit visible light upon irradiation; the amount of light produced is dependent on stopping power (closely related to LET) according to Birks’ law. Doping plastic scintillators can be used to tune their sensitivity to specific LET ranges. Methods: GEANT4.10.1 Monte Carlo (MC) was used to evaluate the response of various scintillator dopant combinations. MC radiation transport and scintillator light response were validated against previously published literature. Current work involves evaluating detector response experimentally; to that end, a detector prototype with interchangeable scintillator housing was constructed. Measurement set-up guides light emitted by the scintillator to a photomultiplier tube via a glass taper junction coupled to an optical fiber. The resulting signal is measured by an electrometer, and normalized to dose readout from a diode. Measurements have been done using clinical electron and orthovoltage beams. MC response (simulated scintillator light normalized to dose scored inside the scintillating volume) was evaluated for four different LET radiations for an undoped and 1%Pb doped scintillator (σ=0.85%). Simulated incident electrons included: 0.05, 0.1, 0.2, 6, 12, and 18 MeV; these energies correspond to a range of stopping power (related to LET) values ranging from 1.824 to 11.09 MeVcm{sup 2}g{sup −1} (SCOL from NIST-ESTAR). Results: Initial MC results show a distinct divergence in scintillator response as LET increases. The response for undoped plastic scintillator indicated a 35.0% increase in signal when going from 18 MeV (low LET) to 0.05 MeV (high LET) while 1%-Pb doped scintillator indicated a 100.9% increase. Conclusion: After validating MC against measurement, simulations will be used to test various concentrations (2%, 4%, 6%) of

  2. Radiation dosimeter built with plastic scintillator

    International Nuclear Information System (INIS)

    Barrea, R.A.

    1990-01-01

    Ionization chambers, with air equivalent plastic walls, have been used as the main x-ray dosimetry system, since its response permits to give the doses in Roentgens. From the commercial availability of plastic scintillators with air equivalent atomic number we have studied its use in x-ray dosimetry. This paper devised a system with which it is possible to reduce side effects that introduce errors and obtain a response independent of energy, from 20 to 180 KeV and it is also expected to behave that way for higher energies. The system has a high efficiency and precision for a wide dose range and it is then a new alternative to measure x-ray doses. Results obtained with a prototype, built ad-hog, make it possible its use with dosimetric purposes with several advantages over conventional. (author)

  3. First-principles Electronic Structure Calculations for Scintillation Phosphor Nuclear Detector Materials

    Science.gov (United States)

    Canning, Andrew

    2013-03-01

    Inorganic scintillation phosphors (scintillators) are extensively employed as radiation detector materials in many fields of applied and fundamental research such as medical imaging, high energy physics, astrophysics, oil exploration and nuclear materials detection for homeland security and other applications. The ideal scintillator for gamma ray detection must have exceptional performance in terms of stopping power, luminosity, proportionality, speed, and cost. Recently, trivalent lanthanide dopants such as Ce and Eu have received greater attention for fast and bright scintillators as the optical 5d to 4f transition is relatively fast. However, crystal growth and production costs remain challenging for these new materials so there is still a need for new higher performing scintillators that meet the needs of the different application areas. First principles calculations can provide a useful insight into the chemical and electronic properties of such materials and hence can aid in the search for better new scintillators. In the past there has been little first-principles work done on scintillator materials in part because it means modeling f electrons in lanthanides as well as complex excited state and scattering processes. In this talk I will give an overview of the scintillation process and show how first-principles calculations can be applied to such systems to gain a better understanding of the physics involved. I will also present work on a high-throughput first principles approach to select new scintillator materials for fabrication as well as present more detailed calculations to study trapping process etc. that can limit their brightness. This work in collaboration with experimental groups has lead to the discovery of some new bright scintillators. Work supported by the U.S. Department of Homeland Security and carried out under U.S. Department of Energy Contract no. DE-AC02-05CH11231 at Lawrence Berkeley National Laboratory.

  4. Development of metallic molds for the large volume plastic scintillator fabrication

    International Nuclear Information System (INIS)

    Calvo, Wilson A.P.; Vieira, Jose M.; Rela, Paulo R.; Bruzinga, Wilson A.; Araujo, Eduardo P.; Costa Junior, Nelson P.; Hamada, Margarida M.

    1997-01-01

    The plastic scintillators are radiation detectors made of organic fluorescent compounds dissolved in a solidified polymer matrix. The manufacturing process of large volume detectors (55 liters) at low cost, by polymerization of the styrene monomer plus PPO and POPOP scintillators, was studied in this paper. Metallic molds of ASTM 1200 aluminum and AISI 304 stainless steel were produced by TIG welding process since the polymerization reaction is very exothermic. The measurements of transmittance, luminescence, X-ray fluorescence and light output were carried out in the plastic scintillators made using different metallic molds. The characterization results of the detectors produced in an open system using ASTM 1200 aluminum mold show that there is not quality change in the scintillator, even with aluminum being considered as unstable for styrene monomer. Therefore, the ASTM 1200 aluminum was found to be the best alternative to produce the detector by an open system polymerization. (author). 11 refs., 8 figs., 1 tab

  5. Research of radiation firmness of transparent melamine-formaldehyde polymers

    International Nuclear Information System (INIS)

    Lebedev, V.V.

    2007-01-01

    Radiation properties of the transparent melamine-formaldehyde polymers offered in quality polymeric basis for making of plastic scintillators are explored in this work. Plastic scintillator is composition, that consists of polymer (polymeric basis) and organic fluorescent addition. Scintillation efficiency and light output are basic properties of plastic scintillators. Firmness to influencing of ionizing radiation is important property of scintillators. From all types of scintillators the plastic are most radiation-proof. Cured melamine-formaldehyde resin and melamine-formaldehyde resin modified by different polyol modifiers was a research object. It is shown that radiation firmness for given types of polymeric material considerably depends on composition of polymer and from technology and temperature condition of its receipt. By the method IR-spectroscopy the structural changes in melamine-formaldehyde polymers under action of irradiation were explored. The maximal falling after the irradiation was marked in intensity of luminescence, which went down to 50% from an initial level. Like the coefficients of admission for all compositions got worse of a to 30-35% level from initial one. Mechanical properties went down on 20-30%. The radiation loss of mass made less than 1% for all polymers. With the increase of temperature of curing firmness rises. Thus, on the basis of the conducted researches radiation firmness for different melamine-formaldehyde polymers is determined and processes what is going on in material under action of radiation are studied. The limited doses of irradiation for each of explored polymers are determined. (authors)

  6. submitter Preparation and luminescence properties of ZnO:Ga – polystyrene composite scintillator

    CERN Document Server

    Burešová, Hana; Turtos, Rosana Martinez; Jarý, Vítězslav; Mihóková, Eva; Beitlerová, Alena; Pjatkan, Radek; Gundacker, Stefan; Auffray, Etiennette; Lecoq, Paul; Nikl, Martin; Čuba, Václav

    2016-01-01

    Highly luminescent ZnO:Ga-polystyrene composite (ZnO:Ga-PS) with ultrafast subnanosecond decay was prepared by homogeneous embedding the ZnO:Ga scintillating powder into the scintillating organic matrix. The powder was prepared by photo-induced precipitation with subsequent calcination in air and Ar/H2 atmospheres. The composite was subsequently prepared by mixing the ZnO:Ga powder into the polystyrene (10 wt% fraction of ZnO:Ga) and press compacted to the 1 mm thick pellet. Luminescent spectral and kinetic characteristics of ZnO:Ga were preserved. Radioluminescence spectra corresponded purely to the ZnO:Ga scintillating phase and emission of polystyrene at 300-350 nm was absent. These features suggest the presence of non-radiative energy transfer from polystyrene host towards the ZnO:Ga scintillating phase which is confirmed by the measurement of X-ray excited scintillation decay with picosecond time resolution. It shows an ultrafast rise time below the time resolution of the experiment (18 ps) and a single-...

  7. Liquid scintillation systems and apparatus for measuring high-energy radiation emitted by samples in standard laboratory test tubes

    International Nuclear Information System (INIS)

    Benvenutti, R.A.

    1976-01-01

    Liquid scintillation detection system employs improved sample holders in which the cap of a glass vial is provided with a well for receiving a standard laboratory test tube containing a radioactive sample. The well is immersed in a liquid scintillator in the vial, the scintillator containing lead acetate solution to enhance its efficiency. A commercially available beta-counting liquid scintillation apparatus is modified to provide gamma-counting with the improved sample holders

  8. Test of radiation hardness of pcCVD detectors

    Energy Technology Data Exchange (ETDEWEB)

    Schlemme, Steffen [GSI Helmholtzzentrum fuer Schwerionenforschung, Darmstadt (Germany); Technische Universitaet Darmstadt (Germany); Enders, Joachim [Technische Universitaet Darmstadt (Germany); Figuera, P.; Salamone, S. [LNS-INFN Catania (Italy); Fruehauf, J.; Kis, Mladen; Kratz, A.; Kurz, N.; Loechner, S.; Nociforo, Chiara; Schirru, Fabio; Szczepanczyk, B.; Traeger, M.; Visinka, R. [GSI Helmholtzzentrum fuer Schwerionenforschung, Darmstadt (Germany); Musumarra, A. [LNS-INFN Catania (Italy); University of Catania (Italy)

    2016-07-01

    The new in-flight separator Super-FRS is under construction at the Facility for Antiproton and Ion Research (FAIR, Darmstadt). Ion rates up to 3 x 10{sup 11} {sup 238}U/spill demand an adaption of detectors to a high radiation environment. A test experiment to investigate the radiation hardness of polycrystalline diamond detectors (pcCVD) was performed at the LNS-INFN in Catania using a {sup 12}C beam at 62 MeV/u and intensities of up to 1.5 pnA. The setup consisted of pcCVD strip detectors to measure the beam profile, a single crystal diamond detector to calibrate the ionisation chamber working in current mode as a beam intensity monitor and a pcCVD sample to be irradiated. The IC used was designed for FAIR and showed a stable counting rate allowing us to calibrate and perform beam intensity measurements with it. The total measured counts on the sample were 8.25 x 10{sup 11} counts/mm{sup 2} over a period of 60 hours. Digital waveforms of the pcCVD signals were taken with an oscilloscope and analysed. The results showed no change of the pcCVD signal properties during the entire irradiation.

  9. Investigation on neutron/gamma discrimination phenomena in plastic scintillators

    International Nuclear Information System (INIS)

    Blanc, Pauline

    2014-01-01

    This PhD topic was born from misunderstandings and incomplete knowledge of the mechanism and relative effectiveness of neutron and gamma-ray (n/γ) discrimination between plastic scintillators compared to liquid scintillators. The shape of the light pulse these materials generate following interaction with an ionizing particle (predominantly recoil protons in the case of neutrons and electrons in the case of gamma-rays) is different in time in a way that depends on the detected particle (nature and energy). It is this fact that enables separation (PSD). The behavior in liquid scintillators has been extensively studied experimentally for practical applications. Only recently has it been shown that a weak separation can also be achieved using specially prepared plastics. The study of this system presents an open field and the understanding of both liquids and plastics with respect to their PSD properties is far from complete. This work is dedicated to exploring the fundamental photophysical phenomena at play in the generation of luminescence emission, following the interaction of ionizing radiation with organic scintillators. For this purpose, firstly a detailed literature review of the state-of-the-art has been conducted extending from 1960 to the present day. Secondly a complete characterization of the main scintillating materials has been conducted to define their fluorescence properties and the characteristics of their scintillation under irradiation. Thirdly a proton beam has been used to simulate recoil protons to quantify under controlled laboratory conditions their specific energy deposition in a plastic scintillator with PSD properties. The fourth part of this thesis is devoted to the study of PSD efficiency of scintillators as a function of their molecular structure. This investigation has led to a plastic scintillator prepared in our laboratory with good PSD properties and a patent submission. Finally, photophysical experiments were performed using a

  10. Scintillators module for the detection of X or gamma rays, realization process, and detector set up incorporating such modules

    International Nuclear Information System (INIS)

    Colomb, Gilbert; Delattre, Dominique; Ricodeau, Jean; Roziere, Guy.

    1981-01-01

    The invention concerns a scintillator module (sodium iodide) for the detection of X and γ rays by light conversion into electrical signals. This conversion is made through optical elements associated to the scintillator and receiving the light generated by X and γ rays. This module is constituted of elementary scintillators separated by interstices, arranged on a transparent support exposed to incident radiations. These interstices constitute optical discontinuities at the surface between two elementary and successive scintillators. The interstices are blocked up by polymerisable and transparent resin to the light of the scintillator, and charged by magnesia powder reflecting this light [fr

  11. Electrical properties and radiation hardness of SOI systems with multilayer buried dielectric

    International Nuclear Information System (INIS)

    Barchuk, I.P.; Kilchitskaya, V.I.; Lysenko, V.S.

    1997-01-01

    In this work SOI structures with buried SiO 2 -Si 3 N 4 -SiO 2 layers have been fabricated by the ZMR-technique with the aim of improving the total dose radiation hardness of the buried dielectric layer. To optimize the fabrication process, buried layers were investigated by secondary ion mass spectrometry before and after the ZMR process, and the obtained results were compared with electrical measurements. It is shown that optimization of the preparation processes of the initial buried dielectric layers provides ZMR SOI structures with multilayer buried isolation, which are of high quality for both Si film interfaces. Particular attention is paid to the investigation of radiation-induced charge trapping in buried insulators. Buried isolation structures with a nitride layer exhibit significant reduction of radiation-induced positive charge as compared to classical buried SiO 2 layers produced by either the ZMR or the SIMOX technique

  12. Evaluation of NaI(TL) and plastic scintillators for use in remote, unattended, and portal monitoring

    International Nuclear Information System (INIS)

    Staples, P.; Audia, J.; Bai, Y.; Briggs, M.; Halbig, J.K.; Ianakiev, K.D.

    1998-01-01

    The authors have evaluated and compared some of the relevant operating characteristics of NaI and plastic scintillators for use in various safeguards monitoring applications. These include a sensitivity analysis of the two scintillators to various radiation fields and scintillator response as affected by environmental temperature. A comparison of experiment and modeling via the Monte Carlo N-Particle (MCNP) code has been performed to validate the calculational techniques. This then enables complex detector situations to be simulated with increased confidence

  13. Gammastic: towards a pseudo-gamma spectrometry in plastic scintillators

    International Nuclear Information System (INIS)

    Hamel, Matthieu; Dehe-Pittance, Chrystele; Coulon, Romain; Carrel, Frederick; Pillot, Philippe; Barat, Eric; Dautremer, Thomas; Montagu, Thierry; Normand, Stephane

    2013-06-01

    War against CBRN-E threats needs to continuously develop sensors with improved detection efficiency. More particularly, this topic concerns the NR controls for homeland security. A first analysis requires indeed a fast gamma spectrometry so as to detect potential special nuclear materials (SNM). To this aim, plastic scintillators could represent the best alternative for the production of large-scale, low-cost radiation portal monitors to be deployed on boarders, tolls, etc. Although they are known to be highly sensitive to gamma rays, due to their poor resolution, information relative to the nature of the SNM is tricky. Thus, only the Compton edge is obtained after interaction, and no information of the photoelectric peak is observed. This project concerns new developments on a possible pseudo-gamma spectrometry performed with plastic scintillators. This project is articulated on a combination of two developments: - The design of new materials most suitable for recovering the photoelectric peak after gamma interaction with the scintillator. This work concerns mainly plastic scintillators loading with heavy elements, such as lead or bismuth. - The analysis of the resulting signal with smart algorithms. This work is thus a pluri-disciplinary work performed at CEA LIST and embeds 4 main disciplines: MCNPX simulations (simulated spectra), chemistry of materials (preparation of various plastic scintillators with different properties), instrumentation (lab experiments) and smart algorithms. Really impressive results were obtained with the unfolding of simulated spectra at various energies (from 241 Am to 60 Co) and an innovative approach was proposed to counter-balance the quenching effect of luminescence by heavy elements in plastic scintillators. (authors)

  14. Preparation of plastic scintillator detectors and physicochemical parameter characterization

    International Nuclear Information System (INIS)

    Hamada, M.M.; Mesquita, C.H. de.

    1988-10-01

    The development of plastic scintillators for use in the nuclear radiation detection is described. The detectors were fabricated by the polymerization of styrene with organic fluors. The organic fluors used were PPO (1,4 diphenyl-oxazol) and POPOP 1,4-di-2-(5-phenyl-oxazolil) - benzene in proportions of 0,5 and 0,05% respectively. Physicochemical parameters related to the quality of this detector are investigated at this laboratory. The evaluation of its fluorescence characteristics, density, melting softening, refractive index, molecular weight, gamma and alpha spectrometry characteristics and finally the comparative pulse height analysis indicate that the plastic scintillator produced at this laboratory is comparable with others already described. (author)

  15. Alkali metal and alkali earth metal gadolinium halide scintillators

    Science.gov (United States)

    Bourret-Courchesne, Edith; Derenzo, Stephen E.; Parms, Shameka; Porter-Chapman, Yetta D.; Wiggins, Latoria K.

    2016-08-02

    The present invention provides for a composition comprising an inorganic scintillator comprising a gadolinium halide, optionally cerium-doped, having the formula A.sub.nGdX.sub.m:Ce; wherein A is nothing, an alkali metal, such as Li or Na, or an alkali earth metal, such as Ba; X is F, Br, Cl, or I; n is an integer from 1 to 2; m is an integer from 4 to 7; and the molar percent of cerium is 0% to 100%. The gadolinium halides or alkali earth metal gadolinium halides are scintillators and produce a bright luminescence upon irradiation by a suitable radiation.

  16. Radiation damage in BaF2 crystals

    International Nuclear Information System (INIS)

    Woody, C.L.; Kierstead, J.A.; Levy, P.W.; Stoll, S.

    1991-01-01

    The effects of radiation damage and recovery have been studied in BaF 2 crystals exposed to 60 Co radiation. The change in optical transmission and scintillation light output have been measured as a function of dose up to 4.7 x 10 6 rad. Although some crystals exhibit a small change in transmission, a greater change in scintillation light output is observed. Several 25 cm long crystals whichhave been irradiated show large changes in both transmission and light output. Recovery from radiation damage has been studied as a function of time and exposure to UV light. A long lived radiation induced phosphorescence has been observed in all irradiated samples which is distinct from the standard fast and slow scintillation emissions. The emission spectrum of the phosphorescence has been measured and shown a peakat ∼330 nm, near the region of the slow scintillation component. Results are given on the dependence of the decay time of the phosphorescence with dose

  17. Systems for detecting and recording hard corpuscular and electromagnetic radiations using a superconducting element

    International Nuclear Information System (INIS)

    Drukier, A.K.; Valette, Claude; Waysand, Georges.

    1975-01-01

    The invention relates to systems for detecting hard X ray or gamma radiations above 5 keV in energy, intended, for example, for gammagraphy by emission in the biological, anatomic and medical fields. It describes systems of the above type which directly give the image of a radiation distribution, that is to say without intermediate processing. Another purpose of the invention is to provide the devices with main memory measuring the radiation distribution, in other words systems that display the said data for as long as the operator deems necessary and that can be reset, that is to say returned to measuring conditions, immediately. The invention makes use of the properties of type I superconductors [fr

  18. Luminescence and scintillation properties of YAG:Ce single crystal and optical ceramics

    CERN Document Server

    Mihóková, E; Mareš, J A; Beitlerová, A; Vedda, A; Nejezchleb, K; Blažek, K; D’Ambrosio, C

    2007-01-01

    We use various techniques to study optical and scintillation properties of Ce-doped yttrium aluminum garnet, Y3Al5O12 (YAG:Ce), in the form of a high-quality industrial single crystal. This was compared to optical ceramics prepared from YAG:Ce nanopowders. We present experimental data in the areas of optical absorption, radioluminescence, scintillation decay, photoelectron yield, thermally stimulated luminescence and radiation-induced absorption. The results point to an interesting feature—the absence of antisite (YAl, i.e. Y at the Al site) defects in optical ceramics. The scintillation decay of the ceramics is faster than that of the single crystal, but its photoelectron yield (measured with 1 μs integration time) is about 30–40% lower. Apart from the photoelectron yield value the YAG:Ce optical ceramic is fully comparable to a high quality industrial YAG:Ce single crystal and can become a competitive scintillator material.

  19. System and plastic scintillator for discrimination of thermal neutron, fast neutron, and gamma radiation

    Science.gov (United States)

    Zaitseva, Natalia P.; Carman, M. Leslie; Faust, Michelle A.; Glenn, Andrew M.; Martinez, H. Paul; Pawelczak, Iwona A.; Payne, Stephen A.

    2017-05-16

    A scintillator material according to one embodiment includes a polymer matrix; a primary dye in the polymer matrix, the primary dye being a fluorescent dye, the primary dye being present in an amount of 3 wt % or more; and at least one component in the polymer matrix, the component being selected from a group consisting of B, Li, Gd, a B-containing compound, a Li-containing compound and a Gd-containing compound, wherein the scintillator material exhibits an optical response signature for thermal neutrons that is different than an optical response signature for fast neutrons and gamma rays. A system according to one embodiment includes a scintillator material as disclosed herein and a photodetector for detecting the response of the material to fast neutron, thermal neutron and gamma ray irradiation.

  20. Effects of device scaling and geometry on MOS radiation hardness assurance

    International Nuclear Information System (INIS)

    Shaneyfelt, M.R.; Fleetwood, D.M.; Winokur, P.S.; Schwank, J.R.; Meisenheimer, T.L.

    1993-01-01

    In this work the authors investigate the effects of transistor scaling and geometry on radiation hardness. The total dose response is shown to depend strongly on transistor channel length. Specifically, transistors with shorter gate lengths tend to show more negative threshold-voltage shifts during irradiation than transistors with longer gate lengths. Similarly, transistors with longer gate lengths tend to show more positive threshold-voltage shifts during post-irradiation annealing than transistors with shorter gate lengths. These differences in radiation response, caused by differences in transistor size and geometry, will be important to factor into test-structure-to-IC correlations necessary to support cost-effective Qualified Manufacturers List (QML) hardness assurance. Transistors with minimum gate length (more negative ΔV th ) will have a larger effect on standby power supply current for an IC at high dose rates, such as in a weapon environment, where worst-case response is associated with negative threshold-voltage shifts during irradiation. On the other hand, transistors with maximum gate length (more positive ΔV th ) will have a larger effect on the timing parameters of an IC at low dose rates, such as in a space environment, where worst-case response is represented by positive threshold-voltage shifts after postirradiation anneal. The channel size and geometry effects they observe cannot be predicted from simple scaling models, but occur because of real differences in oxide-, interface-, and border-trap charge densities among devices of different sizes

  1. Organic Scintillator Detector Response Simulations with DRiFT

    Energy Technology Data Exchange (ETDEWEB)

    Andrews, Madison Theresa [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Bates, Cameron Russell [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Mckigney, Edward Allen [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Rising, Michael Evan [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Pinilla, Maria Isabel [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Solomon, Jr., Clell Jeffrey [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Sood, Avneet [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2016-12-19

    Accurate detector modeling is a requirement to design systems in many non-proliferation scenarios; by determining a Detector’s Response Function (DRF) to incident radiation, it is possible characterize measurements of unknown sources. DRiFT is intended to post-process MCNP® output and create realistic detector spectra. Capabilities currently under development include the simulation of semiconductor, gas, and (as is discussed in this work) scintillator detector physics. Energy spectra and pulse shape discrimination (PSD) trends for incident photon and neutron radiation have been reproduced by DRiFT.

  2. RD50 Status Report 2008 - Radiation hard semiconductor devices for very high luminosity colliders

    CERN Document Server

    Balbuena, Juan Pablo; Campabadal, Francesca; Díez, Sergio; Fleta, Celeste; Lozano, Manuel; Pellegrini, Giulio; Rafí, Joan Marc; Ullán, Miguel; Creanza, Donato; De Palma, Mauro; Fedele, Francesca; Manna, Norman; Kierstead, Jim; Li, Zheng; Buda, Manuela; Lazanu, Sorina; Pintilie, Lucian; Pintilie, Ioana; Popa, Andreia-Ioana; Lazanu, Ionel; Collins, Paula; Fahrer, Manuel; Glaser, Maurice; Joram, Christian; Kaska, Katharina; La Rosa, Alessandro; Mekki, Julien; Moll, Michael; Pacifico, Nicola; Pernegger, Heinz; Goessling, Claus; Klingenberg, Reiner; Weber, Jens; Wunstorf, Renate; Roeder, Ralf; Stolze, Dieter; Uebersee, Hartmut; Cihangir, Selcuk; Kwan, Simon; Spiegel, Leonard; Tan, Ping; Bruzzi, Mara; Focardi, Ettore; Menichelli, David; Scaringella, Monica; Breindl, Michael; Eckert, Simon; Köhler, Michael; Kuehn, Susanne; Parzefall, Ulrich; Wiik, Liv; Bates, Richard; Blue, Andrew; Buttar, Craig; Doherty, Freddie; Eklund, Lars; Bates, Alison G; Haddad, Lina; Houston, Sarah; James, Grant; Mathieson, Keith; Melone, J; OShea, Val; Parkes, Chris; Pennicard, David; Buhmann, Peter; Eckstein, Doris; Fretwurst, Eckhart; Hönniger, Frank; Khomenkov, Vladimir; Klanner, Robert; Lindström, Gunnar; Pein, Uwe; Srivastava, Ajay; Härkönen, Jaakko; Lassila-Perini, Katri; Luukka, Panja; Mäenpää, Teppo; Tuominen, Eija; Tuovinen, Esa; Eremin, Vladimir; Ilyashenko, Igor; Ivanov, Alexandr; Kalinina, Evgenia; Lebedev, Alexander; Strokan, Nikita; Verbitskaya, Elena; Barcz, Adam; Brzozowski, Andrzej; Kaminski, Pawel; Kozlowski, Roman; Kozubal, Michal; Luczynski, Zygmunt; Pawlowski, Marius; Surma, Barbara; Zelazko, Jaroslaw; de Boer, Wim; Dierlamm, Alexander; Frey, Martin; Hartmann, Frank; Zhukov, Valery; Barabash, L; Dolgolenko, A; Groza, A; Karpenko, A; Khivrich, V; Lastovetsky, V; Litovchenko, P; Polivtsev, L; Campbell, Duncan; Chilingarov, Alexandre; Fox, Harald; Hughes, Gareth; Jones, Brian Keith; Sloan, Terence; Samadashvili, Nino; Tuuva, Tuure; Affolder, Anthony; Allport, Phillip; Bowcock, Themis; Casse, Gianluigi; Vossebeld, Joost; Cindro, Vladimir; Dolenc, Irena; Kramberger, Gregor; Mandic, Igor; Mikuž, Marko; Zavrtanik, Marko; Zontar, Dejan; Gil, Eduardo Cortina; Grégoire, Ghislain; Lemaitre, Vincent; Militaru, Otilia; Piotrzkowski, Krzysztof; Kazuchits, Nikolai; Makarenko, Leonid; Charron, Sébastien; Genest, Marie-Helene; Houdayer, Alain; Lebel, Celine; Leroy, Claude; Aleev, Andrey; Golubev, Alexander; Grigoriev, Eugene; Karpov, Aleksey; Martemianov, Alxander; Rogozhkin, Sergey; Zaluzhny, Alexandre; Andricek, Ladislav; Beimforde, Michael; Macchiolo, Anna; Moser, Hans-Günther; Nisius, Richard; Richter, Rainer; Gorelov, Igor; Hoeferkamp, Martin; Metcalfe, Jessica; Seidel, Sally; Toms, Konstantin; Hartjes, Fred; Koffeman, Els; van der Graaf, Harry; Visschers, Jan; Kuznetsov, Andrej; Sundnes Løvlie, Lars; Monakhov, Edouard; Svensson, Bengt G; Bisello, Dario; Candelori, Andrea; Litovchenko, Alexei; Pantano, Devis; Rando, Riccardo; Bilei, Gian Mario; Passeri, Daniele; Petasecca, Marco; Pignatel, Giorgio Umberto; Bernardini, Jacopo; Borrello, Laura; Dutta, Suchandra; Fiori, Francesco; Messineo, Alberto; Bohm, Jan; Mikestikova, Marcela; Popule, Jiri; Sicho, Petr; Tomasek, Michal; Vrba, Vaclav; Broz, Jan; Dolezal, Zdenek; Kodys, Peter; Tsvetkov, Alexej; Wilhelm, Ivan; Chren, Dominik; Horazdovsky, Tomas; Kohout, Zdenek; Pospisil, Stanislav; Solar, Michael; Sopko, Vít; Sopko, Bruno; Uher, Josef; Horisberger, Roland; Radicci, Valeria; Rohe, Tilman; Bolla, Gino; Bortoletto, Daniela; Giolo, Kim; Miyamoto, Jun; Rott, Carsten; Roy, Amitava; Shipsey, Ian; Son, SeungHee; Demina, Regina; Korjenevski, Sergey; Grillo, Alexander; Sadrozinski, Hartmut; Schumm, Bruce; Seiden, Abraham; Spence, Ned; Hansen, Thor-Erik; Artuso, Marina; Borgia, Alessandra; Lefeuvre, Gwenaelle; Guskov, J; Marunko, Sergey; Ruzin, Arie; Tylchin, Tamir; Boscardin, Maurizio; Dalla Betta, Gian - Franco; Gregori, Paolo; Piemonte, Claudio; Ronchin, Sabina; Zen, Mario; Zorzi, Nicola; Garcia, Carmen; Lacasta, Carlos; Marco, Ricardo; Marti i Garcia, Salvador; Minano, Mercedes; Soldevila-Serrano, Urmila; Gaubas, Eugenijus; Kadys, Arunas; Kazukauskas, Vaidotas; Sakalauskas, Stanislavas; Storasta, Jurgis; Vidmantis Vaitkus, Juozas; CERN. Geneva. The LHC experiments Committee; LHCC

    2010-01-01

    The objective of the CERN RD50 Collaboration is the development of radiation hard semiconductor detectors for very high luminosity colliders, particularly to face the requirements of a possible upgrade scenario of the LHC.This document reports the status of research and main results obtained after the sixth year of activity of the collaboration.

  3. A position-sensitive scintillation detector for two-dimensional angular correlation of annihilation radiation using metal-package position-sensitive photomultiplier tubes

    International Nuclear Information System (INIS)

    Inoue, Koji; Nagai, Yasuyoshi; Saito, Haruo; Nagashima, Yasuyuki; Hyodo, Toshio; Muramatsu, Shinichi; Nagai, Shota

    1999-01-01

    We have constructed and tested a prototype of a new position sensitive γ-ray detector which consists of an array of 2.6x2.6x18 mm 3 BGO scintillator blocks, a light guide, and four metal-package position-sensitive photomultiplier tubes (R5900-00-C8) recently developed by Hamamatsu Photonics Co. Ltd. Scalability of the detector of this type makes it possible to construct a larger detector using many PS-PMTs, which will be useful for the two-dimensional angular correlation of annihilation radiation apparatus

  4. Synthesis of plastic scintillation microspheres: Evaluation of scintillators

    International Nuclear Information System (INIS)

    Santiago, L.M.; Bagán, H.; Tarancón, A.; Garcia, J.F.

    2013-01-01

    The use of plastic scintillation microspheres (PSm) appear to be an alternative to liquid scintillation for the quantification of alpha and beta emitters because it does not generate mixed wastes after the measurement (organic and radioactive). In addition to routine radionuclide determinations, PSm can be used for further applications, e.g. for usage in a continuous monitoring equipment, for measurements of samples with a high salt concentration and for an extractive scintillation support which permits the separation, pre-concentration and measurement of the radionuclides without additional steps of elution and sample preparation. However, only a few manufacturers provide PSm, and the low number of regular suppliers reduces its availability and restricts the compositions and sizes available. In this article, a synthesis method based on the extraction/evaporation methodology has been developed and successfully used for the synthesis of plastic scintillation microspheres. Seven different compositions of plastic scintillation microspheres have been synthesised; PSm1 with polystyrene, PSm2 with 2,5-Diphenyloxazol(PPO), PSm3 with p-terphenyl (pT), PSm4 with PPO and 1,4-bis(5-phenyloxazol-2-yl) (POPOP), PSm5 pT and (1,4-bis [2-methylstyryl] benzene) (Bis-MSB), PSm6 with PPO, POPOP and naphthalene and PSm7 with pT, Bis-MSB and naphthalene. The synthesised plastic scintillation microspheres have been characterised in terms of their morphology, detection capabilities and alpha/beta separation capacity. The microspheres had a median diameter of approximately 130 μm. Maximum detection efficiency values were obtained for the PSm4 composition as follows 1.18% for 3 H, 51.2% for 14 C, 180.6% for 90 Sr/ 90 Y and 76.7% for 241 Am. Values of the SQP(E) parameter were approximately 790 for PSm4 and PSm5. These values show that the synthesised PSm exhibit good scintillation properties and that the spectra are at channel numbers higher than in commercial PSm. Finally, the addition

  5. Indirect X-ray Detectors Based on Inkjet-Printed Photodetectors with a Screen-Printed Scintillator Layer.

    Science.gov (United States)

    Oliveira, Juliana; Correia, Vitor; Sowade, Enrico; Etxebarria, Ikerne; Rodriguez, Raul D; Mitra, Kalyan Y; Baumann, Reinhard R; Lanceros-Mendez, Senentxu

    2018-04-18

    Organic photodetectors (PDs) based on printing technologies will allow to expand the current field of PD applications toward large-area and flexible applications in areas such as medical imaging, security, and quality control, among others. Inkjet printing is a powerful digital tool for the deposition of smart and functional materials on various substrates, allowing the development of electronic devices such as PDs on various substrates. In this work, inkjet-printed PD arrays, based on the organic thin-film transistor architecture, have been developed and applied for the indirect detection of X-ray radiation using a scintillator ink as an X-ray absorber. The >90% increase of the photocurrent of the PDs under X-ray radiation, from about 53 nA without the scintillator film to about 102 nA with the scintillator located on top of the PD, proves the suitability of the developed printed device for X-ray detection applications.

  6. Feasibility study of the real-time IMRT dosimetry using a scintillation screen

    Energy Technology Data Exchange (ETDEWEB)

    Lim, Sang Wook; Yi, Byong Yong; Ko, Young Eun [Asan Medical Center, College of Medicine, University of Ulsan, Seoul (Korea, Republic of)] (and others)

    2004-03-15

    To study the feasibility of verifying real-time 2-D dose distribution measurement system with the scintillation screen for the quality assurance. The water phantom consisted of a scintillation screen (LANEX fast screen, Kodak, USA) that was axially located in the middle of an acrylic cylinder with a diameter of 25 cm. The charge-coupled device (CCD) camera was attached to the phantom in order to capture the visible light from the scintillation screen. To observe the dose distribution in real time, the intensity of the light from the scintillator was converted to a dosage. The isodose contours of the calculations from RTP and those of the measurements using the scintillation screen were compared for the arc therapy and the intensity modulated radiation therapy (IMRT). The kernel, expressed as a multiplication of two error functions, was obtained in order to correct the sensitivity of the CCD of the camera and the scintillation screen. When comparing the calculated isodose and measured isodose, a discrepancy of less than 8 mm in the high dose region was observed. Using the 2-D dosimetry system, the relationship between the light and the dosage could be found, and real-time verification of the dose distribution was feasible.

  7. Feasibility study of the real-time IMRT dosimetry using a scintillation screen

    International Nuclear Information System (INIS)

    Lim, Sang Wook; Yi, Byong Yong; Ko, Young Eun

    2004-01-01

    To study the feasibility of verifying real-time 2-D dose distribution measurement system with the scintillation screen for the quality assurance. The water phantom consisted of a scintillation screen (LANEX fast screen, Kodak, USA) that was axially located in the middle of an acrylic cylinder with a diameter of 25 cm. The charge-coupled device (CCD) camera was attached to the phantom in order to capture the visible light from the scintillation screen. To observe the dose distribution in real time, the intensity of the light from the scintillator was converted to a dosage. The isodose contours of the calculations from RTP and those of the measurements using the scintillation screen were compared for the arc therapy and the intensity modulated radiation therapy (IMRT). The kernel, expressed as a multiplication of two error functions, was obtained in order to correct the sensitivity of the CCD of the camera and the scintillation screen. When comparing the calculated isodose and measured isodose, a discrepancy of less than 8 mm in the high dose region was observed. Using the 2-D dosimetry system, the relationship between the light and the dosage could be found, and real-time verification of the dose distribution was feasible

  8. Cherenkov radiation effects on counting efficiency in extremely quenched liquid scintillation samples

    International Nuclear Information System (INIS)

    Grau Carles, A.; Grau Malonda, A.; Rodriguez Barquero, L.

    1993-01-01

    The CIEMAT/NIST tracer method has successfully standardized nuclides with diverse quench values and decay schemes in liquid scintillation counting. However, the counting efficiency is computed inaccurately for extremely quenched samples. This article shows that when samples are extremely quenched, the counting efficiency in high-energy beta-ray nuclides depends principally on the Cherenkov effect. A new technique is described for quench determination, which makes the measurement of counting efficiency possible when scintillation counting approaches zero. A new efficiency computation model for pure beta-ray nuclides is also described. The results of the model are tested experimentally for 89 Sr, 90 Y, 36 Cl and 204 Tl nuclides with independence of the quench level. (orig.)

  9. Design and test of a scintillation dosimeter for dosimetry measurements of high energy radiotherapy beams; Conception et realisation d'un dosimetre a scintillation adapte a la dosimetrie de faisceaux de rayonnements ionisants en faisceaux de rayonnements ionisants en radiotherapie

    Energy Technology Data Exchange (ETDEWEB)

    Fontbonne, J.M

    2002-12-01

    This work describes the design and evaluation of the performances of a scintillation dosimeter developed for the dosimetry of radiation beams used in radiotherapy. The dosimeter consists in a small plastic scintillator producing light which is guided by means of a plastic optical fiber towards photodetectors. In addition to scintillation, high energy ionizing radiations produce Cerenkov light both in the scintillator and the optical fiber. Based on a wavelength analysis, we have developed a deconvolution technique to measure the scintillation light in the presence of Cerenkov light. We stress the advantages that are anticipated from plastic scintillator, in particular concerning tissue or water equivalence (mass stopping power, mass attenuation or mass energy absorption coefficients). We show that detectors based on this material have better characteristics than conventional dosimeters such as ionisation chambers or silicon detectors. The deconvolution technique is exposed, as well as the calibration procedure using an ionisation chamber. We have studied the uncertainty of our dosimeter. The electronics noise, the fiber transmission, the deconvolution technique and the calibration errors give an overall combined experimental uncertainty of about 0,5%. The absolute response of the dosimeter is studied by means of depth dose measurements. We show that absolute uncertainty with photons or electrons beams with energies ranging from 4 MeV to 25 MeV is less than {+-} 1 %. Last, at variance with other devices, our scintillation dosimeter does not need dose correction with depth. (author)

  10. Practical use of a plastic scintillator for quality assurance of electron beam therapy.

    Science.gov (United States)

    Yogo, Katsunori; Tatsuno, Yuya; Tsuneda, Masato; Aono, Yuki; Mochizuki, Daiki; Fujisawa, Yoshiki; Matsushita, Akihiro; Ishigami, Minoru; Ishiyama, Hiromichi; Hayakawa, Kazushige

    2017-06-07

    Quality assurance (QA) of clinical electron beams is essential for performing accurate and safe radiation therapy. However, with advances in radiation therapy, QA has become increasingly labor-intensive and time-consuming. In this paper, we propose a tissue-equivalent plastic scintillator for quick and easy QA of clinical electron beams. The proposed tool comprises a plastic scintillator plate and a charge-coupled device camera that enable the scintillation light by electron beams to be recorded with high sensitivity and high spatial resolution. Further, the Cerenkov image is directly subtracted from the scintillation image to discriminate Cerenkov emissions and accurately measure the dose profiles of electron beams with high spatial resolution. Compared with conventional methods, discrepancies in the depth profile improved from 7% to 2% in the buildup region via subtractive corrections. Further, the output brightness showed good linearity with dose, good reproducibility (deviations below 1%), and dose rate independence (within 0.5%). The depth of 50% dose measured with the tool, an index of electron beam quality, was within  ±0.5 mm of that obtained with an ionization chamber. Lateral brightness profiles agreed with the lateral dose profiles to within 4% and no significant improvement was obtained using Cerenkov corrections. Field size agreed to within 0.5 mm with those obtained with ionization chamber. For clinical QA of electron boost treatment, a disk scintillator that mimics the shape of a patient's breast is applied. The brightness distribution and dose, calculated using a treatment planning system, was generally acceptable for clinical use, except in limited zones. Overall, the proposed plastic scintillator plate tool efficiently performs QA for electron beam therapy and enables simultaneous verification of output constancy, beam quality, depth, and lateral dose profiles during monthly QAs at lower doses of irradiation (small monitor units, MUs).

  11. Development of Quantum Devices and Algorithms for Radiation Detection and Radiation Signal Processing

    International Nuclear Information System (INIS)

    El Tokhy, M.E.S.M.E.S.

    2012-01-01

    The main functions of spectroscopy system are signal detection, filtering and amplification, pileup detection and recovery, dead time correction, amplitude analysis and energy spectrum analysis. Safeguards isotopic measurements require the best spectrometer systems with excellent resolution, stability, efficiency and throughput. However, the resolution and throughput, which depend mainly on the detector, amplifier and the analog-to-digital converter (ADC), can still be improved. These modules have been in continuous development and improvement. For this reason we are interested with both the development of quantum detectors and efficient algorithms of the digital processing measurement. Therefore, the main objective of this thesis is concentrated on both 1. Study quantum dot (QD) devices behaviors under gamma radiation 2. Development of efficient algorithms for handling problems of gamma-ray spectroscopy For gamma radiation detection, a detailed study of nanotechnology QD sources and infrared photodetectors (QDIP) for gamma radiation detection is introduced. There are two different types of quantum scintillator detectors, which dominate the area of ionizing radiation measurements. These detectors are QD scintillator detectors and QDIP scintillator detectors. By comparison with traditional systems, quantum systems have less mass, require less volume, and consume less power. These factors are increasing the need for efficient detector for gamma-ray applications such as gamma-ray spectroscopy. Consequently, the nanocomposite materials based on semiconductor quantum dots has potential for radiation detection via scintillation was demonstrated in the literature. Therefore, this thesis presents a theoretical analysis for the characteristics of QD sources and infrared photodetectors (QDIPs). A model of QD sources under incident gamma radiation detection is developed. A novel methodology is introduced to characterize the effect of gamma radiation on QD devices. The rate

  12. Comprehensive renal scintillation procedures in spinal cord injury: comparison with excretory urography

    International Nuclear Information System (INIS)

    Lloyd, L.K.; Dubovsky, E.V.; Bueschen, A.J.; Witten, D.M.; Scott, J.W.; Kuhlemeier, K.; Stover, S.L.

    1981-01-01

    A 131 iodine orthoiodohippurate comprehensive renal scintillation procedure was performed and compared to results of excretory urography in 200 spinal cord injury patients. No severe urographic abnormalities were undetected by the comprehensive renal scintillation procedure. Only 1.4 per cent of renal units had greater than minimal pyelocaliectasis or ureterectasis in the presence of a normal radionuclide examination. A relatively large number of abnormalities were detected on the renal scintillation procedure when the excretory urogram was normal. Serial followup will be required to determine the significance of these findings but present data suggest that a comprehensive renal scintillation procedure and a plain film of the kidneys, ureters and bladder may be used for screening upper urinary tract abnormalities in lieu of an excretory urogram. This is particularly advantageous for the spinal cord injury population, since there have been no toxic or allergic reactions reported, no bowel preparation or dehydration is required and there is relatively low radiation exposure

  13. Interference of dissolved salts in Cerenkov and liquid scintillation estimation of 90Sr

    International Nuclear Information System (INIS)

    Pulhani, Vandana; Jha, S.K.; Tripathi, R.M.; Reddy, Priyanka; Bhade, Sonali

    2014-01-01

    Quenching is the most important effect occurring in Cerenkov and LSC because it affects the efficiency of conversion of β particles into light. Bore well water samples are very often concentrated by evaporation to reduce the detection limit which can also increase the dissolved solid content (TDS) in the sample. Some ground waters are inherently having higher TDS. Self-absorption of beta-particle radiation by the sample especially the lower-energy beta particles depends on sample thickness and density. Environmental samples, after applying the radiochemical procedure, are also estimated by Cerenkov/LSC and might be affected by colour quenching. To get best measurements using Liquid Scintillation and Cerenkov radiations, it is necessary to avoid high salt concentrations and colors which may weaken energy transfers within scintillator cocktails and sample medium. Therefore the degree of self-absorption and quench should be evaluated and taken into account in the calibration. Efficiency is represented as a function spectral quench parameter of external standard SQP(E). The quenching effect of dissolved solids on the efficiency of estimation of 90 Sr by Cerenkov and Liquid Scintillation are studied

  14. Radiation-hard/high-speed parallel optical links

    International Nuclear Information System (INIS)

    Gan, K.K.; Buchholz, P.; Kagan, H.P.; Kass, R.D.; Moore, J.; Smith, D.S.; Wiese, A.; Ziolkowski, M.

    2014-01-01

    We have designed an ASIC for use in a parallel optical engine for a new layer of the ATLAS pixel detector in the initial phase of the LHC luminosity upgrade. The ASIC is a 12-channel VCSEL (Vertical Cavity Surface Emitting Laser) array driver capable of operating up to 5 Gb/s per channel. The ASIC is designed using a 130 nm CMOS process to enhance the radiation-hardness. A scheme for redundancy has also been implemented to allow bypassing of a broken VCSEL. The ASIC also contains a power-on reset circuit that sets the ASIC to a default configuration with no signal steering. In addition, the bias and modulation currents of the individual channels are programmable. The performance of the first prototype ASIC up to 5 Gb/s is satisfactory. Furthermore, we are able to program the bias and modulation currents and to bypass a broken VCSEL channel. We are currently upgrading our design to allow operation at 10 Gb/s per channel yielding an aggregated bandwidth of 120 Gb/s. Some preliminary results of the design will be presented

  15. Pulse-resolved radiotherapy dosimetry using fiber-coupled organic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Ravnsborg Beierholm, A.

    2011-05-15

    This PhD project pertains to the development and adaptation of a dosimetry system that can be used to verify the delivery of radiation in modern radiotherapy modalities involving small radiation fields and dynamic radiation delivery. The dosimetry system is based on fibre-coupled organic scintillators and can be perceived as a well characterized, independent alternative to the methods that are in clinical use today. The dosimeter itself does not require a voltage supply, and is composed of water equivalent materials. The dosimeter can be fabricated with a sensitive volume smaller than a cubic millimeter, which is small enough to resolve the small radiation fields encountered in modern radiotherapy. The fast readout of the dosimeter enables measurements on the same time scale as the pulsed radiation delivery from the medical linear accelerators used for treatment. The dosimetry system, comprising fiber-coupled organic scintillators and data acquisition hardware, was developed at the Radiation Research Division at Risoe DTU and tested using clinical x-ray beams at hospitals in Denmark and abroad. Measurements of output factors and percentage depth dose were performed and compared with reference values and Monte Carlo simulations for static square radiation fields for standard (4 cm x 4 cm to 20 cm x 20 cm) and small (down to 0.6 cm x 0.6 cm) field sizes. The accuracy of most of the obtained measurements was good, agreeing with reference and simulated dose values to within 2 % standard deviation for both standard and small fields. This thesis concludes that the new pulse-resolved dosimetry system holds great potential for modern radiotherapy applications, such as stereotactic radiotherapy and intensity-modulated radiotherapy. (Author)

  16. Pulse-resolved radiotherapy dosimetry using fiber-coupled organic scintillators

    International Nuclear Information System (INIS)

    Ravnsborg Beierholm, A.

    2011-05-01

    This PhD project pertains to the development and adaptation of a dosimetry system that can be used to verify the delivery of radiation in modern radiotherapy modalities involving small radiation fields and dynamic radiation delivery. The dosimetry system is based on fibre-coupled organic scintillators and can be perceived as a well characterized, independent alternative to the methods that are in clinical use today. The dosimeter itself does not require a voltage supply, and is composed of water equivalent materials. The dosimeter can be fabricated with a sensitive volume smaller than a cubic millimeter, which is small enough to resolve the small radiation fields encountered in modern radiotherapy. The fast readout of the dosimeter enables measurements on the same time scale as the pulsed radiation delivery from the medical linear accelerators used for treatment. The dosimetry system, comprising fiber-coupled organic scintillators and data acquisition hardware, was developed at the Radiation Research Division at Risoe DTU and tested using clinical x-ray beams at hospitals in Denmark and abroad. Measurements of output factors and percentage depth dose were performed and compared with reference values and Monte Carlo simulations for static square radiation fields for standard (4 cm x 4 cm to 20 cm x 20 cm) and small (down to 0.6 cm x 0.6 cm) field sizes. The accuracy of most of the obtained measurements was good, agreeing with reference and simulated dose values to within 2 % standard deviation for both standard and small fields. This thesis concludes that the new pulse-resolved dosimetry system holds great potential for modern radiotherapy applications, such as stereotactic radiotherapy and intensity-modulated radiotherapy. (Author)

  17. An organic dye in a polymer matrix – A search for a scintillator with long luminescent lifetime

    DEFF Research Database (Denmark)

    Lindvold, Lars René; Beierholm, Anders Ravnsborg; Andersen, Claus Erik

    2010-01-01

    Fiber-coupled organic plastic scintillators enable dose rate monitoring in conjunction with pulsed radiation sources like linear medical accelerators. The accelerator, however, generates a significant amount of stray ionizing radiation. This radiation excites the long optical fiber cable (15–20 m...

  18. Scintillation crystals for positron emission tomography having a non reflecting band

    International Nuclear Information System (INIS)

    Thompson, C.J.

    1992-01-01

    This invention relates generally to positron emission tomography, a sub-field of the class of medical imaging techniques using ionizing radiation and image reconstruction techniques; and more particularly to devices which use an array of scintillation detectors to detect the annihilation radiation from positron disintegration and use this information to reconstruct an image of the distribution of positron emitting isotope within a body section. 6 figs

  19. Neutron flux measurement with 6Li and 7Li dual glass scintillators by γ compensation method

    International Nuclear Information System (INIS)

    Ji Changsong; Zhang Shulan; Zhang Shuheng

    1996-01-01

    Based on the characteristics of 6 Li glass scintillator which is sensitive to both neutron and gamma rays, and 7 Li glass scintillator which is sensitive to gamma rays only, a new method of detecting weak neutron flux under interference of strong gamma radiation has been investigated by means of 6 Li- 7 Li pair glass scintillator gamma compensation method. The result of neutron flux measurement by above-mentioned method with an error of about 1% when the gamma ray interference is up to 18.7% has been obtained

  20. Neutron flux measurement with 6Li and 7Li dual glass scintillators by γ compensation method

    International Nuclear Information System (INIS)

    Ji Changsong; Zhang Shulan; Zhang Shuheng

    1998-01-01

    Based on the characteristics of 6 Li glass scintillator which is sensitive to both neutron and gamma rays, and 7 Li glass scintillator which is sensitive to gamma rays only, a new method of detecting weak neutron flux under interference of strong gamma radiation has been investigated by mans of 6 Li- 7 Li dual glass scintillator gamma compensation method. The result of neutron flux measurement by above-mentioned method with an error of about 1% when the gamma ray interference is up to 18.7% has been obtained

  1. Evaluating scintillators used in radiation detectors of medical imaging systems by the effective fidelity index method

    International Nuclear Information System (INIS)

    Kandarakis, I.; Cavouras, D.; Prassopoulos, P.; Kanellopoulos, E.; Nomicos, C.D.; Panayiotakis, G.S.

    1999-01-01

    Objective: The performance of medical X-ray image receptors depends: (1) on the scintillator light emission efficiency; and (2) on the compatibility of the scintillator light spectrum with the spectral sensitivity of the light detector (film, photocathode, or photodiode), employed in conjunction with the scintillator. In this study, a scintillator performance measure, the effective fidelity index (EFI), is defined as function of both the scintillator light emission efficiency and spectral compatibility. Materials and Method: CsI:Na, Gd 2 O 2 S:Tb and La 2 O 2 S:Tb scintillators were employed in the form of phosphor screens prepared in our laboratory with various coating thicknesses. The screens were irradiated with X-rays employing tube voltages ranging between 50-120 kVp. Results: The EFI performance of CsI:Na was found to increase with screen coating thickness and it was best when combined with the orthochromatic film or the ES/20 photocathode. Gd 2 O 2 S:Tb showed peak EFI performance at 70 mg/cm 2 coating thickness and it was well combined with the light detectors considered. Conclusion: In accordance with our results, CsI:Na may be employed in radiography when adequately protected against humidity. Gd 2 O 2 S:Tb suitability for conventional imaging was verified and it was found that it may be useful in all types of digital imaging. La 2 O 2 S:Tb could also be used in digital detectors of imaging applications demanding medium X-ray tube voltages

  2. Effect of geomagnetic storms on VHF scintillations observed at low latitude

    Science.gov (United States)

    Singh, S. B.; Patel, Kalpana; Singh, A. K.

    2018-06-01

    A geomagnetic storm affects the dynamics and composition of the ionosphere and also offers an excellent opportunity to study the plasma dynamics. In the present study, we have used the VHF scintillations data recorded at low latitude Indian station Varanasi (Geomag. latitude = 14^{°}55^' }N, long. = 154^{°}E) which is radiated at 250 MHz from geostationary satellite UFO-02 during the period 2011-2012 to investigate the effects of geomagnetic storms on VHF scintillation. Various geomagnetic and solar indices such as Dst index, Kp index, IMF Bz and solar wind velocity (Vx) are used to describe the geomagnetic field variation observed during geomagnetic storm periods. These indices are very helpful to find out the proper investigation and possible interrelation between geomagnetic storms and observed VHF scintillation. The pre-midnight scintillation is sometimes observed when the main phase of geomagnetic storm corresponds to the pre-midnight period. It is observed that for geomagnetic storms for which the recovery phase starts post-midnight, the probability of occurrence of irregularities is enhanced during this time and extends to early morning hours.

  3. Liquid scintillator for 2D dosimetry for high-energy photon beams

    International Nuclear Information System (INIS)

    Poenisch, Falk; Archambault, Louis; Briere, Tina Marie; Sahoo, Narayan; Mohan, Radhe; Beddar, Sam; Gillin, Michael T.

    2009-01-01

    Complex radiation therapy techniques require dosimetric verification of treatment planning and delivery. The authors investigated a liquid scintillator (LS) system for application for real-time high-energy photon beam dosimetry. The system was comprised of a transparent acrylic tank filled with liquid scintillating material, an opaque outer tank, and a CCD camera. A series of images was acquired when the tank with liquid scintillator was irradiated with a 6 MV photon beam, and the light data measured with the CCD camera were filtered to correct for scattering of the optical light inside the liquid scintillator. Depth-dose and lateral profiles as well as two-dimensional (2D) dose distributions were found to agree with results from the treatment planning system. Further, the corrected light output was found to be linear with dose, dose rate independent, and is robust for single or multiple acquisitions. The short time needed for image acquisition and processing could make this system ideal for fast verification of the beam characteristics of the treatment machine. This new detector system shows a potential usefulness of the LS for 2D QA.

  4. Liquid scintillator for 2D dosimetry for high-energy photon beams

    Energy Technology Data Exchange (ETDEWEB)

    Poenisch, Falk; Archambault, Louis; Briere, Tina Marie; Sahoo, Narayan; Mohan, Radhe; Beddar, Sam; Gillin, Michael T. [Department of Radiation Physics, University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Boulevard., Unit 94, Houston, Texas 77030 (United States)

    2009-05-15

    Complex radiation therapy techniques require dosimetric verification of treatment planning and delivery. The authors investigated a liquid scintillator (LS) system for application for real-time high-energy photon beam dosimetry. The system was comprised of a transparent acrylic tank filled with liquid scintillating material, an opaque outer tank, and a CCD camera. A series of images was acquired when the tank with liquid scintillator was irradiated with a 6 MV photon beam, and the light data measured with the CCD camera were filtered to correct for scattering of the optical light inside the liquid scintillator. Depth-dose and lateral profiles as well as two-dimensional (2D) dose distributions were found to agree with results from the treatment planning system. Further, the corrected light output was found to be linear with dose, dose rate independent, and is robust for single or multiple acquisitions. The short time needed for image acquisition and processing could make this system ideal for fast verification of the beam characteristics of the treatment machine. This new detector system shows a potential usefulness of the LS for 2D QA.

  5. RD50 Status Report 2009/2010 - Radiation hard semiconductor devices for very high luminosity colliders

    CERN Document Server

    Moll, Michael

    2012-01-01

    The objective of the CERN RD50 Collaboration is the development of radiation hard semiconductor detectors for very high luminosity colliders, particularly to face the requirements for the upgrade of the LHC detectors. This document reports on the status of research and main results obtained in the years 2009 and 2010.

  6. Event localization in bulk scintillator crystals using coded apertures

    Energy Technology Data Exchange (ETDEWEB)

    Ziock, K.P. [Oak Ridge National Laboratory, Oak Ridge, TN (United States); Department of Physics and Astronomy, University of Tennessee, Knoxville, TN (United States); Braverman, J.B. [Department of Physics and Astronomy, University of Tennessee, Knoxville, TN (United States); Fabris, L.; Harrison, M.J.; Hornback, D.; Newby, J. [Oak Ridge National Laboratory, Oak Ridge, TN (United States)

    2015-06-01

    The localization of radiation interactions in bulk scintillators is generally limited by the size of the light distribution at the readout surface of the crystal/light-pipe system. By finding the centroid of the light spot, which is typically of order centimeters across, practical single-event localization is limited to ~2 mm/cm of crystal thickness. Similar resolution can also be achieved for the depth of interaction by measuring the size of the light spot. Through the use of near-field coded-aperture techniques applied to the scintillation light, light transport simulations show that for 3-cm-thick crystals, more than a five-fold improvement (millimeter spatial resolution) can be achieved both laterally and in event depth. At the core of the technique is the requirement to resolve the shadow from an optical mask placed in the scintillation light path between the crystal and the readout. In this paper, experimental results are presented that demonstrate the overall concept using a 1D shadow mask, a thin-scintillator crystal and a light pipe of varying thickness to emulate a 2.2-cm-thick crystal. Spatial resolutions of ~1 mm in both depth and transverse to the readout face are obtained over most of the crystal depth.

  7. Event localization in bulk scintillator crystals using coded apertures

    International Nuclear Information System (INIS)

    Ziock, K.P.; Braverman, J.B.; Fabris, L.; Harrison, M.J.; Hornback, D.; Newby, J.

    2015-01-01

    The localization of radiation interactions in bulk scintillators is generally limited by the size of the light distribution at the readout surface of the crystal/light-pipe system. By finding the centroid of the light spot, which is typically of order centimeters across, practical single-event localization is limited to ~2 mm/cm of crystal thickness. Similar resolution can also be achieved for the depth of interaction by measuring the size of the light spot. Through the use of near-field coded-aperture techniques applied to the scintillation light, light transport simulations show that for 3-cm-thick crystals, more than a five-fold improvement (millimeter spatial resolution) can be achieved both laterally and in event depth. At the core of the technique is the requirement to resolve the shadow from an optical mask placed in the scintillation light path between the crystal and the readout. In this paper, experimental results are presented that demonstrate the overall concept using a 1D shadow mask, a thin-scintillator crystal and a light pipe of varying thickness to emulate a 2.2-cm-thick crystal. Spatial resolutions of ~1 mm in both depth and transverse to the readout face are obtained over most of the crystal depth

  8. An Efficient Digital Pulse Shape Discrimination Technique for Scintillation Detectors Based on FPGA

    International Nuclear Information System (INIS)

    Kamel, M.S.

    2014-01-01

    Different techniques for pulse discrimination (PSD) of the scintillation pulses have been developed. The PSD of scintillation pulese can been used in several applications as Positron Emission Topography (PET) system. Each technique analyzes the resulting pulses from the absorption of radiation in the scintillation pulses were filtered and digitized then it is captured using DAQ, and it sent to the host computer for processing. The spatial resolution of images that generated in PET system can be improved by applying the proposed PSD. In this thesis various digital PSD techniques are proposed to discriminate the scintillation pulses. These techniques are based on discrete sine transform (DST). discrete cosine transform (DCT). Discrete hartley transform (DHT), Discrete Goertzel transform (DGT),and principal component analysis (PCA). Then the output coefficients of the discrete transforms are classified using one of the following classifiers T-test,tuned, or support vector machine (SVM).

  9. Basic study on radiation distribution sensing with normal optical fiber

    International Nuclear Information System (INIS)

    Naka, R.; Kawarabayashi, J.; Uritani, A.; Iguchi, T.; Kaneko, J.; Takeuchi, H.; Kakuta, T.

    2000-01-01

    Recently, some methods of radiation distribution sensing with optical fibers have been proposed. These methods employ scintillating fibers or scintillators with wavelength-shifting fibers. The positions of radiation interactions are detected by applying a time-of-flight (TOF) technique to the scintillation photon propagation. In the former method, the attenuation length for the scintillation photons in the scintillating fiber is relatively short, so that the operating length of the sensor is limited to several meters. In the latter method, a radiation distribution cannot continuously be obtained but discretely. To improve these shortcomings, a normal optical fiber made of polymethyl methacrylate (PMMA) is used in this study. Although the scintillation efficiency of PMMA is very low, several photons are emitted through interaction with a radiation. The fiber is transparent for the emitted photons to have a relatively long operating length. A radiation distribution can continuously be obtained. This paper describes a principle of the position sensing method based on the time of flight technique and preliminary results obtained for 90 Sr- 90 Y beta rays, 137 Cs gamma rays, and 14 MeV neutrons. The spatial resolutions for the above three kinds of radiations are 0.30 m, 0.37 m, 0.13 m, and the detection efficiencies are 1.1 x 10 -3 , 1.6 x 10 -7 , 5.4 x 10 -6 , respectively, with 10 m operation length. The results of a spectroscopic study on the optical property of the fiber are also described. (author)

  10. Radiation measurement

    International Nuclear Information System (INIS)

    Go, Sung Jin; Kim, Seung Guk; No, Gyeong Seok; Park, Myeong Hwan; Ann, Bong Seon

    1998-03-01

    This book explains technical terms about radiation measurement, which are radiation, radiation quantity and unit such as prefix of international unit, unit for defence purposes of radiation, coefficient of radiation and interaction, kinds and principles of radiation detector, ionization chamber, G-M counter, G-M tube, proportional counter, scintillation detector, semiconductor radiation detector, thermoluminescence dosimeter, PLD, others detector, radiation monitor, neutron detector, calibration of radiation detector, statistics of counting value, activation analysis and electronics circuit of radiation detector.

  11. Gaseous photomultipliers for the readout of scintillators and detection Cherenkov radiation

    International Nuclear Information System (INIS)

    Peskov, V.; Borovik-Romanov, A.

    1993-11-01

    The latest achievements in the development of gaseous detectors for registering UV and visible photons are described. Possible modifications of their design for some particular applications such as the readout of crystal scintillators. noble liquids, fibers and for large area Cherenkov detectors are discussed

  12. A mixed analog-digital radiation hard technology for high energy physics electronics DMILL (Durci Mixte sur Isolant Logico-Linéaire)

    CERN Document Server

    Beuville, E; Borgeaud, P; Fourches, N T; Rouger, M; Blanc, J P; Bruel, M; Delevoye-Orsier, E; Gautier, J; Du Port de Pontcharra, J; Truche, R; Dupont-Nivet, E; Flament, O; Leray, J L; Martin, J L; Montaron, J; Borel, G; Brice, J M; Chatagnon, P; Terrier, C; Aubert, Jean-Jacques; Delpierre, P A; Habrard, M C; Potheau, R; CERN. Geneva. Detector Research and Development Committee

    1992-01-01

    The high radiation level expected in the inner regions of the high luminosity LHC detectors (gamma and neutron) will require radiation hardened electronics. A consortium between the CEA (Commissariat a l'Energie Atomique) and Thomson TMS (Thomson Composants Militaires et Spatiaux) has been created to push for the development and the industrialization of a nascent technology which looks particularly adapted to the needs of HEP electronics. This technology, currently under development at the LETI(CEA), uses a SIMOX substrate with an epitaxial silicon film. It includes CMOS, JFETs and vertical bipolar transistors with a potential multi-megarad hardness. The CMOS and bipolar transistors constitute a rad-hard BiCMOS which will be useful to design analog and digital high-speed architectures. JFETs, which have intrinsically high hardness behaviour and low noise performances even at low temperature will enable very rad-hard, low noise front end electronics to be designed. Present results, together with the improvemen...

  13. Radiation hardness of silicon detectors manufactured on wafers from various sources

    International Nuclear Information System (INIS)

    Dezillie, B.; Bates, S.; Glaser, M.; Lemeilleur, F.; Leroy, C.

    1997-01-01

    Impurity concentrations in the initial silicon material are expected to play an important role for the radiation hardness of silicon detectors, during their irradiation and for their evolution with time after irradiation. This work reports on the experimental results obtained with detectors manufactured using various float-zone (FZ) and epitaxial-grown material. Preliminary results comparing the changes in leakage current and full depletion voltage of FZ and epitaxial detectors as a function of fluence and of time after 10 14 cm -2 proton irradiation are given. The measurement of charge collection efficiency for epitaxial detectors is also presented. (orig.)

  14. Hybrid metal organic scintillator materials system and particle detector

    Science.gov (United States)

    Bauer, Christina A.; Allendorf, Mark D.; Doty, F. Patrick; Simmons, Blake A.

    2011-07-26

    We describe the preparation and characterization of two zinc hybrid luminescent structures based on the flexible and emissive linker molecule, trans-(4-R,4'-R') stilbene, where R and R' are mono- or poly-coordinating groups, which retain their luminescence within these solid materials. For example, reaction of trans-4,4'-stilbenedicarboxylic acid and zinc nitrate in the solvent dimethylformamide (DMF) yielded a dense 2-D network featuring zinc in both octahedral and tetrahedral coordination environments connected by trans-stilbene links. Similar reaction in diethylformamide (DEF) at higher temperatures resulted in a porous, 3-D framework structure consisting of two interpenetrating cubic lattices, each featuring basic to zinc carboxylate vertices joined by trans-stilbene, analogous to the isoreticular MOF (IRMOF) series. We demonstrate that the optical properties of both embodiments correlate directly with the local ligand environments observed in the crystal structures. We further demonstrate that these materials produce high luminescent response to proton radiation and high radiation tolerance relative to prior scintillators. These features can be used to create sophisticated scintillating detection sensors.

  15. Rad Hard Active Media For Calorimeters

    CERN Document Server

    Norbeck, E; Möller, A; Onel, Y

    2006-01-01

    Zero-degree calorimeters have limited space and extreme levels of radiation. A simple, low cost, radiation hard design uses tungstenmetal as the absorber and a suitable liquid as the ˇCerenkov radiator. In other applications a PPAC (Parallel Plate Avalanche Counter) operatingwith a suitable atmosphericpressure gas is an attractive active material for a calorimeter. It can be made radiation hard and has sufficient gain in the gas that no electronic components are needed near the detector. It works well even with the highest concentration of shower particles. For this pressure range, R134A (used in auto air conditioners) has many desirable features.

  16. R&D on scintillation materials for novel ionizing radiation detectors for High Energy Physics, medical imaging and industrial applications

    CERN Multimedia

    Chipaux, R; Rinaldi, D; Boursier, Y M; Vasilyev, A; Tikhomirov, V; Morel, C; Choi, Y; Tamulaitis, G

    2002-01-01

    The Crystal Clear Collaboration (CCC) was approved by the Detector R&D Committee as RD18 in 1990 with the objective of developing new inorganic scintillators suitable for crystal electromagnetic calorimeters of LHC experiments. From 1990 to 1994, CCC made an intensive investigation for the quest of the most adequate ideal scintillator for the LHC; three main candidates were identified and extensively studied : CeF$_{3}$, PbWO$_{4}$ and heavy scintillating glasses. Lead tungstate was chosen by CMS and ALICE as the most cost effective crystal compliant to LHC conditions. Today 76648 PWO crystals are installed in CMS and 17920 in ALICE. After this success Crystal clear has continued its investigation on new scintillators and the understanding of scintillation mechanisms and light transfer properties in particular : The understanding of cerium ion as activator, The development of LuAP, LuYAP crystals for medical imaging applications, (CERN patent) Investigation of Ytterbium based scintillators for solar ne...

  17. A theoretical model evaluating the angular distribution of luminescence emission in X-ray scintillating screens

    International Nuclear Information System (INIS)

    Kandarakis, I.; Cavouras, D.; Nikolopoulos, D.; Episkopakis, A.; Kalivas, N.; Liaparinos, P.; Valais, I.; Kagadis, G.; Kourkoutas, K.; Sianoudis, I.; Dimitropoulos, N.; Nomicos, C.; Panayiotakis, G.

    2006-01-01

    The aim of this study was to examine the angular distribution of the light emitted from radiation-excited scintillators in medical imaging detectors. This distribution diverges from Lambert's cosine law and affects the light emission efficiency of scintillators, hence it also affects the dose burden to the patient. In the present study, the angular distribution was theoretically modeled and was used to fit experimental data on various scintillator materials. Results of calculations revealed that the angular distribution is more directional than that predicted by Lambert's law. Divergence from this law is more pronounced for high values of light attenuation coefficient and thick scintillator layers (screens). This type of divergence reduces light emission efficiency and hence it increases the incident X-ray flux required for a given level of image brightness

  18. Scintillation hodoscopes on the basis of hodoscopic photomultipliers using scintillation fibers

    International Nuclear Information System (INIS)

    Alimova, T.V.; Vasil'chenko, V.G.; Vechkanov, G.N.

    1986-01-01

    Scintillation hodoscopes characteristics and their design features have been considered. The space resolution for hodoscopes consisting of 4 layers of scintillation fibres 200 mm long and 1 mm in diameter is 0.4-0.6 mm. With 2 fibres layer 1 m long and 3.8 mm in diameter the space resolution 3 mm has been obtained. A possibility to construct 0.1 mm resolution scintillation hodoscopes is discussed

  19. Performance ratio hardness characteristics polystyrene-metal composite materials

    International Nuclear Information System (INIS)

    Klepikov, V.F.; Prokhorenko, E.M.; Lytvynenko, V.V.; Zakharchenko, A.A.; Hazhmuradov, M.A.

    2015-01-01

    The methods of measuring the hardness of layered polystyrene-metallic composite materials. It is proposed to use powder-like tungsten and powder-like steel as radiation-protective layer. A measurement of the hardness of composites of different composition, and given its dependence on the particle size and their form. The possibility of increasing the hardness of the composites reinforced with metallic additives. Radiation-protective characteristics were calculated for the studied species of composite materials. Influence of the quantitative composition of the metal components is studied on the change of the absorbed dose of gamma radiation

  20. Iterative Monte Carlo simulation with the Compton kinematics-based GEB in a plastic scintillation detector

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Chankyu; Kim, Yewon [Department of Nuclear and Quantum Engineering, KAIST, Daejeon 305-701 (Korea, Republic of); Moon, Myungkook [Neutron Instrumentation Division, KAERI, Daejeon 305-353 (Korea, Republic of); Cho, Gyuseong, E-mail: gscho@kaist.ac.kr [Department of Nuclear and Quantum Engineering, KAIST, Daejeon 305-701 (Korea, Republic of)

    2015-09-21

    Plastic scintillators have been used for gamma ray detection in the fields of dosimetry and homeland security because of their desired characteristics such as a fast decay time, a low production cost, availability in a large-scale, and a tissue-equivalence. Gaussian energy broadening (GEB) in MCNP simulation is an effective treatment for tallies to calculate the broadened response function of a detector similarly to measured spectra. The full width at half maximum (FWHM) of a photopeak has been generally used to compute input parameters required for the GEB treatment. However, it is hard to find the photopeak in measured gamma spectra with plastic scintillators so that computation of the input parameters for the GEB has to be taken with another way. In this study, an iterative method for the GEB treated MCNP simulation to calculate the response function of a plastic scintillator is suggested. Instead of the photopeak, Compton maximum and Compton edge were used to estimate energy broadening in the measured spectra and to determine the GEB parameters. In a demonstration with a CsI(Tl) scintillator, the proposed iterative simulation showed the similar gamma spectra to the existing method using photopeaks. The proposed method was then applied to a polystyrene scintillator, and the simulation result were in agreement with the measured spectra with only a little iteration.

  1. Temperature quenching in LAB based liquid scintillator

    Energy Technology Data Exchange (ETDEWEB)

    Soerensen, A.; Zuber, K. [Technische Universitaet Dresden, Institute for Nuclear- and Particle Physics, Dresden (Germany); Hans, S.; Yeh, M. [Brookhaven National Laboratory, Chemistry Devision, Upton, NY (United States); Junghans, A.R.; Koegler, T.; Wagner, A. [Helmholtz-Zentrum Dresden-Rossendorf, Dresden (Germany); Krosigk, B. v. [Technische Universitaet Dresden, Institute for Nuclear- and Particle Physics, Dresden (Germany); University of British Columbia, Department of Physics and Astronomy, Vancouver, BC (Canada); Lozza, V. [Technische Universitaet Dresden, Institute for Nuclear- and Particle Physics, Dresden (Germany); Laboratorio de Instrumentacao e Fisica Experimental de Particulas, Lisboa (Portugal)

    2018-01-15

    The effect of temperature changes on the light output of LAB based liquid scintillator is investigated in a range from -5 to 30 C with α-particles and electrons in a small scale setup. Two PMTs observe the scintillator liquid inside a cylindrically shaped aluminum cuvette that is heated or cooled and the temperature dependent PMT sensitivity is monitored and corrected. The α-emitting isotopes in dissolved radon gas and in natural Samarium (bound to a LAB solution) excite the liquid scintillator mixtures and changes in light output with temperature variation are observed by fitting light output spectra. Furthermore, also changes in light output by compton electrons, which are generated from external calibration γ-ray sources, is analysed with varying temperature. Assuming a linear behaviour, a combined negative temperature coefficient of (-0.29 ± 0.01)%/ C is found. Considering hints for a particle type dependency, electrons show (-0.17 ± 0.02)%/ C, whereas the temperature dependency seems stronger for α-particles, with (-0.35 ± 0.03)%/ C. Due to a high sampling rate, a pulse shape analysis can be performed and shows an enhanced slow decay component at lower temperatures, pointing to reduced non-radiative triplet state de-excitations. (orig.)

  2. Influence of transfer gate design and bias on the radiation hardness of pinned photodiode CMOS image sensors

    International Nuclear Information System (INIS)

    Goiffon, V.; Estribeau, M.; Cervantes, P.; Molina, R.; Magnan, P.; Gaillardin, M.

    2014-01-01

    The effects of Cobalt 60 gamma-ray irradiation on pinned photodiode (PPD) CMOS image sensors (CIS) are investigated by comparing the total ionizing dose (TID) response of several transfer gate (TG) and PPD designs manufactured using a 180 nm CIS process. The TID induced variations of charge transfer efficiency (CTE), pinning voltage, equilibrium full well capacity (EFWC), full well capacity (FWC) and dark current measured on the different pixel designs lead to the conclusion that only three degradation sources are responsible for all the observed radiation effects: the pre-metal dielectric (PMD) positive trapped charge, the TG sidewall spacer positive trapped charge and, with less influence, the TG channel shallow trench isolation (STI) trapped charge. The different FWC evolutions with TID presented here are in very good agreement with a recently proposed analytical model. This work also demonstrates that the peripheral STI is not responsible for the observed degradations and thus that the enclosed layout TG design does not improve the radiation hardness of PPD CIS. The results of this study also lead to the conclusion that the TG OFF voltage bias during irradiation has no influence on the radiation effects. Alternative design and process solutions to improve the radiation hardness of PPD CIS are discussed. (authors)

  3. Optical and scintillation properties of Sr7%:Ce15%:GdF.sub.3./sub. single crystal

    Czech Academy of Sciences Publication Activity Database

    Fukabori, A.; Kamada, K.; Yanagida, T.; Chani, V.; Aoki, K.; Yokota, Y.; Maeo, S.; Nikl, Martin; Yoshikawa, A.

    2011-01-01

    Roč. 318, č. 1 (2011), s. 1175-1178 ISSN 0022-0248. [International Conference on Crystal Growth (ICCG16) /16./ and International Conference on Vapor Growth and Epitaxy (ICVGE14) /14./. Beijing, 08.08.2010-13.08.2010] Institutional research plan: CEZ:AV0Z10100521 Keywords : radiation * inorganic compounds * scintillator materials * scintillators Subject RIV: BM - Solid Matter Physics ; Magnetism Impact factor: 1.726, year: 2011

  4. Accelerated Aging Test for Plastic Scintillator Gamma Ray Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Kouzes, Richard T. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States)

    2017-05-12

    Polyvinyl toluene (PVT) and polystyrene (PS), collectively referred to as “plastic scintillator,” are synthetic polymer materials used to detect gamma radiation, and are commonly used in instrumentation. Recent studies have revealed that plastic scintillator undergoes an environmentally related material degradation that adversely affects performance under certain conditions and histories. A significant decrease in gamma ray sensitivity has been seen in some detectors in systems as they age. The degradation of sensitivity of plastic scintillator over time is due to a variety of factors, and the term “aging” is used to encompass all factors. Some plastic scintillator samples show no aging effects (no significant change in sensitivity over more than 10 years), while others show severe aging (significant change in sensitivity in less than 5 years). Aging effects arise from weather (variations in heat and humidity), chemical exposure, mechanical stress, light exposure, and loss of volatile components. The damage produced by these various causes can be cumulative, causing observable damage to increase over time. Damage may be reversible up to some point, but becomes permanent under some conditions. It has been demonstrated that exposure of plastic scintillator in an environmental chamber to 30 days of high temperature and humidity (90% relative humidity and 55°C) followed by a single cycle to cold temperature (-30°C) will produce severe fogging in all PVT samples. This thermal cycle will be referred to as the “Accelerated Aging Test.” This document describes the procedure for performing this Accelerated Aging Test.

  5. Nanophosphor composite scintillators comprising a polymer matrix

    Science.gov (United States)

    Muenchausen, Ross Edward; Mckigney, Edward Allen; Gilbertson, Robert David

    2010-11-16

    An improved nanophosphor composite comprises surface modified nanophosphor particles in a solid matrix. The nanophosphor particle surface is modified with an organic ligand, or by covalently bonding a polymeric or polymeric precursor material. The surface modified nanophosphor particle is essentially charge neutral, thereby preventing agglomeration of the nanophosphor particles during formation of the composite material. The improved nanophosphor composite may be used in any conventional scintillator application, including in a radiation detector.

  6. Uncooled Radiation Hard Large Area SiC X-ray and EUV Detectors and 2D Arrays, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — This project seeks to design, fabricate, characterize and commercialize large area, uncooled and radiative hard 4H-SiC EUV ? soft X-ray detectors capable of ultra...

  7. Is the holy grail plastic? Radiation identification from plastic scintillators

    International Nuclear Information System (INIS)

    Butchins, L. J. C.; Gosling, J. M.; Hogbin, M. R. W.; Jones, D. C.; Lacey, R. J.; Stearn, J. G.

    2009-01-01

    Thousands of shipping containers containing Naturally Occurring Radioactive Materials (NORM) made from ceramics, stoneware and other natural products are transported worldwide on a daily basis. Some of these NORM loads are sufficiently radioactive to trigger alarms from plastic scintillator detectors which have limited ability to also identify the radionuclides present thus necessitating secondary inspection which increases the operational overhead. Previous studies have been carried out to ascertain if radionuclide discrimination using plastic scintillators is possible with a variety of approaches including deconvolution and computer learning. In this paper, a two stage algorithm is described. An example implementation of the algorithm is presented, applied to operational data, and has been installed in real time operation on a polyvinyl-toluene (PVT) detector. The approach requires the collection of a large library of spectra using examples of the detectors to be deployed. In this study, data from both actual freight loads passing through a port and predefined freight containing various radionuclides were collected. The library represents freight loads that may contain industrial, medical, nuclear, and NORM radionuclides. The radionuclides in the predefined freight were placed in various orientations and in various amounts of shielding to mimic many different scenarios. Preliminary results on an initial subset of data containing industrial and NORM sources show the number of mis-classifications to be less than 1% of the total test data. Good initial results were obtained even for low energy radionuclides such as 241 Am. Where discrimination is not possible, and principle components overlap, this region or 'cloud' of the n-dimensional plot can be put aside. Those spectra that fall in the 'cloud' can be regarded as suspect and in these cases, some secondary screening will still be necessary. It is predicted that the algorithm will enable recognition of NORM loads

  8. The DosiMap, a new 2D scintillating dosimeter for IMRT quality assurance: Characterization of two Cerenkov discrimination methods

    International Nuclear Information System (INIS)

    Frelin, A-M.; Fontbonne, J-M.; Ban, G.; Colin, J.; Labalme, M.; Batalla, A.; Vela, A.; Boher, P.; Braud, M.; Leroux, T.

    2008-01-01

    New radiation therapy techniques such as IMRT present significant efficiency due to their highly conformal dose distributions. A consequence of the complexity of their dose distributions (high gradients, small irradiation fields, low dose distribution, ...) is the requirement for better precision quality assurance than in classical radiotherapy in order to compare the conformation of the delivered dose with the planned dose distribution and to guarantee the quality of the treatment. Currently this control is mostly performed by matrices of ionization chambers, diode detectors, dosimetric films, portal imaging, or dosimetric gels. Another approach is scintillation dosimetry, which has been developed in the last 15 years mainly through scintillating fiber devices. Despite having many advantages over other methods it is still at an experimental level for routine dosimetry because the Cerenkov radiation produced under irradiation represents an important stem effect. A new 2D water equivalent scintillating dosimeter, the DosiMap, and two different Cerenkov discrimination methods were developed with the collaboration of the Laboratoire de Physique Corpusculaire of Caen, the Comprehensive Cancer Center Francois Baclesse, and the ELDIM Co., in the frame of the MAESTRO European project. The DosiMap consists of a plastic scintillating sheet placed inside a transparent polystyrene phantom. The light distribution produced under irradiation is recorded by a CCD camera. Our first Cerenkov discrimination technique is subtractive. It uses a chessboard pattern placed in front of the scintillator, which provides a background signal containing only Cerenkov light. Our second discrimination technique is colorimetric. It performs a spectral analysis of the light signal, which allows the unfolding of the Cerenkov radiation and the scintillation. Tests were carried out with our DosiMap prototype and the performances of the two discrimination methods were assessed. The comparison of the

  9. The DosiMap, a new 2D scintillating dosimeter for IMRT quality assurance: characterization of two Cerenkov discrimination methods.

    Science.gov (United States)

    Frelin, A M; Fontbonne, J M; Ban, G; Colin, J; Labalme, M; Batalla, A; Vela, A; Boher, P; Braud, M; Leroux, T

    2008-05-01

    New radiation therapy techniques such as IMRT present significant efficiency due to their highly conformal dose distributions. A consequence of the complexity of their dose distributions (high gradients, small irradiation fields, low dose distribution, ...) is the requirement for better precision quality assurance than in classical radiotherapy in order to compare the conformation of the delivered dose with the planned dose distribution and to guarantee the quality of the treatment. Currently this control is mostly performed by matrices of ionization chambers, diode detectors, dosimetric films, portal imaging, or dosimetric gels. Another approach is scintillation dosimetry, which has been developed in the last 15 years mainly through scintillating fiber devices. Despite having many advantages over other methods it is still at an experimental level for routine dosimetry because the Cerenkov radiation produced under irradiation represents an important stem effect. A new 2D water equivalent scintillating dosimeter, the DosiMap, and two different Cerenkov discrimination methods were developed with the collaboration of the Laboratoire de Physique Corpusculaire of Caen, the Comprehensive Cancer Center François Baclesse, and the ELDIM Co., in the frame of the MAESTRO European project. The DosiMap consists of a plastic scintillating sheet placed inside a transparent polystyrene phantom. The light distribution produced under irradiation is recorded by a CCD camera. Our first Cerenkov discrimination technique is subtractive. It uses a chessboard pattern placed in front of the scintillator, which provides a background signal containing only Cerenkov light. Our second discrimination technique is colorimetric. It performs a spectral analysis of the light signal, which allows the unfolding of the Cerenkov radiation and the scintillation. Tests were carried out with our DosiMap prototype and the performances of the two discrimination methods were assessed. The comparison of the

  10. Timing performance of ZnO:Ga nanopowder composite scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Turtos, Rosana M. [Universita degli Studi di Milano-Bicocca, Milano (Italy); Gundacker, Stefan; Lucchini, Marco T.; Lecoq, Paul; Auffray, Etiennette [CERN, Geneva (Switzerland); Prochazkova, Lenka; Cuba, Vaclav [Czech Technical University, Faculty of Nuclear Sciences and Physical Engineering, Prague (Czech Republic); Buresova, Hana [Nuvia a.s, Kralupy nad Vltavou (Czech Republic); Mrazek, Jan [Institute of Photonics and Electronics AS CR, Prague (Czech Republic); Nikl, Martin [Institute of Physics of the AS CR, Prague (Czech Republic)

    2016-11-15

    The implementation of nanocrystal-based composite scintillators as a new generation of ultrafast particle detectors is explored using ZnO:Ga nanopowder. Samples are characterized with a spectral-time resolved photon counting system and pulsed X-rays, followed by coincidence time resolution (CTR) measurements under 511 keV gamma excitation. Results are comparable to CTR values obtained using bulk inorganic scintillators. Bringing the ZnO:Ga nanocrystal's timing performance to radiation detectors could pave the research path towards sub-20 ps time resolution as shown in this contribution. However, an efficiency boost when placing nanopowders in a transparent host constitutes the main challenge in order to benefit from sub-nanosecond recombination times. (copyright 2016 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim)

  11. Radiation detection and measurement

    International Nuclear Information System (INIS)

    Knoll, G.F.

    1979-01-01

    Introductory material covers radiation sources, radiation interactions, general properties of radiation detectors, and counting statistics and error prediction. This is followed by detailed sections on gas-filled detectors, scintillation counters, semiconductor detectors, neutron detectors and spectroscopy, detector electronics and pulse processing, and miscellaneous radiation detectors and applications

  12. Silicon photo-multiplier radiation hardness tests with a beam controlled neutron source

    International Nuclear Information System (INIS)

    Angelone, M.; Pillon, M.; Faccini, R.; Pinci, D.; Baldini, W.; Calabrese, R.; Cibinetto, G.; Cotta Ramusino, A.; Malaguti, R.; Pozzati, M.

    2010-01-01

    Radiation hardness tests were performed at the Frascati Neutron Generator on silicon Photo-Multipliers that were made of semiconductor photon detectors built from a square matrix of avalanche photo-diodes on a silicon substrate. Several samples from different manufacturers have been irradiated, integrating up to 7x10 10 1-MeV-equivalent neutrons per cm 2 . Detector performance was recorded during the neutron irradiation, and a gradual deterioration of their properties began after an integrated fluence of the order of 10 8 1-MeV-equivalent neutrons per cm 2 was reached.

  13. R&D proposal to DRDC fast EM calorimeter with excellent photon angular resolution and energy resolution using scintillating noble liquids

    CERN Document Server

    Chen, M; Sumorok, K; Zhang, X; Gaudreau, M P J; Akimov, D Y; Bolozdynya, A I; Churakov, D; Chernyshov, V; Koutchenkov, A; Kovalenko, A; Kuzichev, V F; Lamkov, V A; Lebedenko, V; Gusev, L; Safronov, G A; Sheinkman, V A; Smirnov, G; Krasnokutsky, R N; Shuvalov, R S; Fedyakin, N N; Sushkov, V V; Akopyan, M V; Gougas, Andreas; Pevsner, A; CERN. Geneva. Detector Research and Development Committee

    1993-01-01

    Recent test beam data have shown fast and large signals for LKr, mixed with >1% LXe. Excellent uniformity in LKr and LXe was achieved over a 37 cm long cell. CsI cathode works well inside LKr/LXe with O(1%) resolution at 5 MeV. Precision calibration in-situ has been demonstrated. Scintillating LKr/LXe detectors are sufficiently radiation hard for LHC environment. These new developments simplify the construction of prototype LKr calorimeter, to demonstrate the superior e/gamma energy resolution and the determination of photon direction using longitudinal and transverse segmentations, which are vital for the detection of the multi-photon states. The constant term in the energy resolution is small, the electronics noise is negligible due to the large signal size. The overall pion/electron suppression is expected to be better than 10-4.

  14. Recent advancements in the development of radiation hard semiconductor detectors for S-LHC

    CERN Document Server

    Fretwurst, E; Al-Ajili, A A; Alfieri, G; Allport, P P; Artuso, M; Assouak, S; Avset, B S; Barabash, L; Barcz, A; Bates, R; Biagi, S F; Bilei, G M; Bisello, D; Blue, A; Blumenau, A; Boisvert, V; Bölla, G; Bondarenko, G B; Borchi, E; Borrello, L; Bortoletto, D; Boscardin, M; Bosisio, L; Bowcock, T J V; Brodbeck, T J; Broz, J; Bruzzi, M; Brzozowski, A; Buda, M; Buhmann, P; Buttar, C; Campabadal, F; Campbell, D; Candelori, A; Casse, G; Cavallini, A; Charron, S; Chilingarov, A G; Chren, D; Cindro, V; Collins, P; Coluccia, R; Contarato, D; Coutinho, J; Creanza, D; Cunningham, L; Dalla Betta, G F; Dawson, I; de Boer, Wim; De Palma, M; Demina, R; Dervan, P; Dittongo, S; Dolezal, Z; Dolgolenko, A; Eberlein, T; Eremin, V; Fall, C; Fasolo, F; Ferbel, T; Fizzotti, F; Fleta, C; Focardi, E; Forton, E; García, C; García-Navarro, J E; Gaubas, E; Genest, M H; Gill, K A; Giolo, K; Glaser, M; Gössling, C; Golovine, V; González-Sevilla, S; Gorelov,I; Goss, J; Gouldwell-Bates, A; Grégoire, G; Gregori, P; Grigoriev, E; Grillo, A A; Groza, A; Guskov, J; Haddad, L; Härkönen, J; Hauler, F; Hoeferkamp, M; Honniger, F; Horazdovsky, T; Horisberger, R P; Horn, M; Houdayer, A; Hourahine, B; Hughes, G; Ilyashenko, Yu S; Irmscher, K; Ivanov, A; Jarasiunas, K; Johansen, K M H; Jones, B K; Jones, R; Joram, C; Jungermann, L; Kalinina, E; Kaminski, P; Karpenko, A; Karpov, A; Kazlauskiene, V; Kazukauskas, V; Khivrich, V; Khomenkov, V P; Kierstead, J A; Klaiber Lodewigs, J M; Klingenberg, R; Kodys, P; Kohout, Z; Korjenevski, S; Koski, M; Kozlowski, R; Kozodaev, M; Kramberger, G; Krasel, O; Kuznetsov, A; Kwan, S; Lagomarsino, S; Lassila-Perini, K M; Lastovetsky, V F; Latino, G; Lazanu, I; Lazanu, S; Lebedev, A; Lebel, C; Leinonen, K; Leroy, C; Li, Z; Lindström, G; Linhart, V; Litovchenko, P G; Litovchenko, A P; Lo Giudice, A; Lozano, M; Luczynski, Z; Luukka, Panja; Macchiolo, A; Makarenko, L F; Mandic, I; Manfredotti, C; Manna, N; Martí i García, S; Marunko, S; Mathieson, K; Melone, J; Menichelli, D; Messineo, A; Metcalfe, J; Miglio, S; Mikuz, M; Miyamoto, J; Moll, M; Monakhov, E; Moscatelli, F; Naoumov, D; Nossarzhevska, E; Nysten, J; Olivero, P; O'Shea, V; Palviainen, T; Paolini, C; Parkes, C; Passeri, D; Pein, U; Pellegrini, G; Perera, L; Petasecca, M; Piemonte, C; Pignatel, G U; Pinho, N; Pintilie, I; Pintilie, L; Polivtsev, L; Polozov, P; Popa, A; Populea, J; Pospísil, S; Pozza, A; Radicci, V; Rafí, J M; Rando, R; Röder, R; Rohe, T; Ronchin, S; Rott, C; Roy, A; Ruzin, A; Sadrozinski, H F W; Sakalauskas, S; Scaringella, M; Schiavulli, L; Schnetzer, S; Schumm, B; Sciortino, S; Scorzoni, A; Segneri, G; Seidela, S; Seiden, A; Sellberg, G; Sellin, P J; Sentenac, D; Shipsey, I; Sícho, P; Sloan, T; Solar, M; Son, S; Sopko, B; Sopko, V; Spencer, N; Stahl, J; Stolze, D; Stone, R; Storasta, J; Strokan, N; Sudzius, M; Surma, B; Suvorov, A; Svensson, B G; Tipton, P; Tomasek, M; Tsvetkov, A; Tuominen, E; Tuovinen, E; Tuuva, T; Tylchin, M; Uebersee, H; Uher, J; Ullán, M; Vaitkus, J V; Velthuis, J; Verbitskaya, E; Vrba, V; Wagner, G; Wilhelm, I; Worm, S; Wright, V; Wunstorf, R; Yiuri, Y; Zabierowski, P; Zaluzhny, A; Zavrtanik, M; Zen, M; Zhukov, V; Zorzi, N

    2005-01-01

    The proposed luminosity upgrade of the Large Hadron Collider (S-LHC) at CERN will demand the innermost layers of the vertex detectors to sustain fluences of about 1016 hadrons/cm2. Due to the high multiplicity of tracks, the required spatial resolution and the extremely harsh radiation field new detector concepts and semiconductor materials have to be explored for a possible solution of this challenge. The CERN RD50 collaboration “Development of Radiation Hard Semiconductor Devices for Very High Luminosity Colliders” has started in 2002 an R&D program for the development of detector technologies that will fulfill the requirements of the S-LHC. Different strategies are followed by RD50 to improve the radiation tolerance. These include the development of defect engineered silicon like Czochralski, epitaxial and oxygen-enriched silicon and of other semiconductor materials like SiC and GaN as well as extensive studies of the microscopic defects responsible for the degradation of irradiated sensors. Furthe...

  15. Testing and simulation of silicon photomultiplier readouts for scintillators in high-energy astronomy and solar physics

    Science.gov (United States)

    Bloser, P. F.; Legere, J. S.; Bancroft, C. M.; Jablonski, L. F.; Wurtz, J. R.; Ertley, C. D.; McConnell, M. L.; Ryan, J. M.

    2014-11-01

    Space-based gamma-ray detectors for high-energy astronomy and solar physics face severe constraints on mass, volume, and power, and must endure harsh launch conditions and operating environments. Historically, such instruments have usually been based on scintillator materials due to their relatively low cost, inherent ruggedness, high stopping power, and radiation hardness. New scintillator materials, such as LaBr3:Ce, feature improved energy and timing performance, making them attractive for future astronomy and solar physics space missions in an era of tightly constrained budgets. Despite this promise, the use of scintillators in space remains constrained by the volume, mass, power, and fragility of the associated light readout device, typically a vacuum photomultiplier tube (PMT). In recent years, silicon photomultipliers (SiPMs) have emerged as promising alternative light readout devices that offer gains and quantum efficiencies similar to those of PMTs, but with greatly reduced mass and volume, high ruggedness, low voltage requirements, and no sensitivity to magnetic fields. In order for SiPMs to replace PMTs in space-based instruments, however, it must be shown that they can provide comparable performance, and that their inherent temperature sensitivity can be corrected for. To this end, we have performed extensive testing and modeling of a small gamma-ray spectrometer composed of a 6 mm×6 mm SiPM coupled to a 6 mm×6 mm ×10 mm LaBr3:Ce crystal. A custom readout board monitors the temperature and adjusts the bias voltage to compensate for gain variations. We record an energy resolution of 5.7% (FWHM) at 662 keV at room temperature. We have also performed simulations of the scintillation process and optical light collection using Geant4, and of the SiPM response using the GosSiP package. The simulated energy resolution is in good agreement with the data from 22 keV to 662 keV. Above ~1 MeV, however, the measured energy resolution is systematically worse than

  16. Testing and simulation of silicon photomultiplier readouts for scintillators in high-energy astronomy and solar physics

    International Nuclear Information System (INIS)

    Bloser, P.F.; Legere, J.S.; Bancroft, C.M.; Jablonski, L.F.; Wurtz, J.R.; Ertley, C.D.; McConnell, M.L.; Ryan, J.M.

    2014-01-01

    Space-based gamma-ray detectors for high-energy astronomy and solar physics face severe constraints on mass, volume, and power, and must endure harsh launch conditions and operating environments. Historically, such instruments have usually been based on scintillator materials due to their relatively low cost, inherent ruggedness, high stopping power, and radiation hardness. New scintillator materials, such as LaBr 3 :Ce, feature improved energy and timing performance, making them attractive for future astronomy and solar physics space missions in an era of tightly constrained budgets. Despite this promise, the use of scintillators in space remains constrained by the volume, mass, power, and fragility of the associated light readout device, typically a vacuum photomultiplier tube (PMT). In recent years, silicon photomultipliers (SiPMs) have emerged as promising alternative light readout devices that offer gains and quantum efficiencies similar to those of PMTs, but with greatly reduced mass and volume, high ruggedness, low voltage requirements, and no sensitivity to magnetic fields. In order for SiPMs to replace PMTs in space-based instruments, however, it must be shown that they can provide comparable performance, and that their inherent temperature sensitivity can be corrected for. To this end, we have performed extensive testing and modeling of a small gamma-ray spectrometer composed of a 6 mm×6 mm SiPM coupled to a 6 mm×6 mm ×10 mm LaBr 3 :Ce crystal. A custom readout board monitors the temperature and adjusts the bias voltage to compensate for gain variations. We record an energy resolution of 5.7% (FWHM) at 662 keV at room temperature. We have also performed simulations of the scintillation process and optical light collection using Geant4, and of the SiPM response using the GosSiP package. The simulated energy resolution is in good agreement with the data from 22 keV to 662 keV. Above ∼1 MeV, however, the measured energy resolution is systematically

  17. Testing and simulation of silicon photomultiplier readouts for scintillators in high-energy astronomy and solar physics

    Energy Technology Data Exchange (ETDEWEB)

    Bloser, P.F., E-mail: Peter.Bloser@unh.edu; Legere, J.S.; Bancroft, C.M.; Jablonski, L.F.; Wurtz, J.R.; Ertley, C.D.; McConnell, M.L.; Ryan, J.M.

    2014-11-01

    Space-based gamma-ray detectors for high-energy astronomy and solar physics face severe constraints on mass, volume, and power, and must endure harsh launch conditions and operating environments. Historically, such instruments have usually been based on scintillator materials due to their relatively low cost, inherent ruggedness, high stopping power, and radiation hardness. New scintillator materials, such as LaBr{sub 3}:Ce, feature improved energy and timing performance, making them attractive for future astronomy and solar physics space missions in an era of tightly constrained budgets. Despite this promise, the use of scintillators in space remains constrained by the volume, mass, power, and fragility of the associated light readout device, typically a vacuum photomultiplier tube (PMT). In recent years, silicon photomultipliers (SiPMs) have emerged as promising alternative light readout devices that offer gains and quantum efficiencies similar to those of PMTs, but with greatly reduced mass and volume, high ruggedness, low voltage requirements, and no sensitivity to magnetic fields. In order for SiPMs to replace PMTs in space-based instruments, however, it must be shown that they can provide comparable performance, and that their inherent temperature sensitivity can be corrected for. To this end, we have performed extensive testing and modeling of a small gamma-ray spectrometer composed of a 6 mm×6 mm SiPM coupled to a 6 mm×6 mm ×10 mm LaBr{sub 3}:Ce crystal. A custom readout board monitors the temperature and adjusts the bias voltage to compensate for gain variations. We record an energy resolution of 5.7% (FWHM) at 662 keV at room temperature. We have also performed simulations of the scintillation process and optical light collection using Geant4, and of the SiPM response using the GosSiP package. The simulated energy resolution is in good agreement with the data from 22 keV to 662 keV. Above ∼1 MeV, however, the measured energy resolution is

  18. Hard X-ray studies on the Castor tokamak

    International Nuclear Information System (INIS)

    Mlynar, J.

    1990-04-01

    The electron runaway processes in tokamaks are discussed with regard to hard X radiation measurements. The origin and confinement of runaway electrons, their bremsstrahlung spectra and the influence of lower hybrid current drive on the distribution of high-energy electrons are analyzed for the case of the Castor tokamak. The hard X-ray spectrometer designed for the Castor tokamak is also described and preliminary qualitative results of hard X-ray measurements are presented. The first series of integral measurements made it possible to map the azimuthal dependence of the hard X radiation

  19. Development of novel UV emitting single crystalline film scintillators

    Science.gov (United States)

    Zorenko, Yu; Gorbenko, V.; Savchyn, V.; Voznyak, T.; Nikl, M.; Mares, J. A.; Martin, T.; Douissard, P.-A.

    2011-04-01

    The work is dedicated to development of new types of UV -emitting scintillators based on single crystalline films (SCF) of aluminimum perovskites and garnets grown by the liquid phase epitaxy (LPE) method. The development of the following three types of UV SCF scintillators is considered in this work: i) Ce-doped SCF of Y-Lu-Al-perovskites with Ce3+ emission in the 360-370 nm range with a decay time of 16-17 ns; ii) Pr-doped SCF of Y-Lu-Al garnets with Pr3+ emission in the 300-400 nm range with a decay time of 13-17 ns; iii) La3+ and Sc3+ doped SCF of Y-Lu-Al-garnets, emitting in the 290-400 nm range due to formation of the LaY,Lu, ScY,Lu and ScAl centers with decay time of 250-575 ns. The results of testing the several novel UV-emitting SCFs scintillators for visualization of X-ray images at ESFR are presented. It is shown that the UV emission of the LuAG:Sc, LuAG:La and LuAG:Pr SCFs is efficient enough for conversion of X-ray to the UV light and that these scintillators can be used for improvement of the resolution of imaging detectors in synchrotron radiation applications.

  20. Development of novel UV emitting single crystalline film scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Zorenko, Yu; Gorbenko, V; Savchyn, V; Voznyak, T [Laboratory of Opoelectronic Materials (LOM), Electronics Department of Ivan Franko National University of Lviv, 79017 Lviv (Ukraine); Nikl, M; Mares, J A [Institute of Physics of ASCR, 162 53 Prague (Czech Republic); Martin, T; Douissard, P-A, E-mail: zorenko@electronics.wups.lviv.ua [ESRF, Instrument Support Group, 6 rue Jules Horoeitz, 38043 Grenoble (France)

    2011-04-01

    The work is dedicated to development of new types of UV -emitting scintillators based on single crystalline films (SCF) of aluminium perovskites and garnets grown by the liquid phase epitaxy (LPE) method. The development of the following three types of UV SCF scintillators is considered in this work: i) Ce-doped SCF of Y-Lu-Al-perovskites with Ce{sup 3+} emission in the 360-370 nm range with a decay time of 16-17 ns; ii) Pr-doped SCF of Y-Lu-Al garnets with Pr{sup 3+} emission in the 300-400 nm range with a decay time of 13-17 ns; iii) La{sup 3+} and Sc{sup 3+} doped SCF of Y-Lu-Al-garnets, emitting in the 290-400 nm range due to formation of the La{sub Y,Lu}, Sc{sub Y,Lu} and Sc{sub Al} centers with decay time of 250-575 ns. The results of testing the several novel UV-emitting SCFs scintillators for visualization of X-ray images at ESFR are presented. It is shown that the UV emission of the LuAG:Sc, LuAG:La and LuAG:Pr SCFs is efficient enough for conversion of X-ray to the UV light and that these scintillators can be used for improvement of the resolution of imaging detectors in synchrotron radiation applications.

  1. Luminescence properties of Ce3+ doped gadolinium-calcium-silicaborate glass scintillator

    International Nuclear Information System (INIS)

    Park, J.M.; Ha, D.H.; Kaewjeang, S.; Maghanemi, U.; Kothan, S.; Kaewkhao, J.; Kim, H.J.

    2016-01-01

    In this work, the Ce 3+ doped gadolinium-calcium-silicaborate glass scintillators of the composition ratio 25Gd 2 O 3 :10CaO:10SiO 2 :(55−x)B 2 O 3 :xCeF 3 , have been fabricated by using the melt-quenching technique. The doping concentration of the Ce 3+ was varied from 0.05 mol% to 2.5 mol%. The 4f-5d transition of the Ce 3+ allowed scintillation with a fast decay time. The absorption spectrum, X-ray induced emission spectrum, photo luminescence spectrum, laser luminescence spectrum and decay time of the scintillators were measured for studying the luminescence properties. From the X-ray induced emission spectrum result, we checked the trend between doping concentration and light yield. The laser induced luminescence spectrum was measured while changing the temperature from 300 K to 10 K. We also measured the decay time by using the laser excitation of the 0.15 mol% Ce 3+ doped glass scintillator. - Highlights: • Ce 3+ doped gadolinium-calcium-silicaborate glass scintillators were developed. • Glass is easily fabricated with large sizes and various doping materials. • The luminescence properties are studied by using various radiation sources. • The light yield and decay time were measured at low temperature. • One decay time component is found.

  2. The development of a high-resolution scintillating fiber tracker with silicon photomultiplier readout

    International Nuclear Information System (INIS)

    Roper Yearwood, Gregorio

    2013-01-01

    In this work I present the design and test results for a novel, modular tracking detector from scintillating fibers which are read out by silicon photomultiplier (SiPM) arrays. The detector modules consist of 0.25 mm thin scintillating fibers which are closely packed in five-layer ribbons. Two ribbons are fixed to both sides of a carbon-fiber composite structure. Custom made SiPM arrays with a photo-detection efficiency of about 50% read out the fibers. Several 860 mm long and 32 mm wide tracker modules were tested in a secondary 12 GeV/c beam at the PS facilities, CERN in November of 2009. During this test a spatial resolution better than 0.05 mm at an average light yield of about 20 photons for a minimum ionizing particle was determined. This work details the characterization of scintillating fibers and silicon photomultipliers of different make and model. It gives an overview of the production of scintillating fiber modules. The behavior of detector modules during the test-beam is analyzed in detail and different options for the front-end electronics are compared. Furthermore, the implementation of the proposed tracking detector from scintillating fibers within the scope of the PERDaix experiment is discussed. The PERDaix detector is a permanent magnet spectrometer with a weight of 40 kg. It consists of 8 tracking detector layers from scintillating fibers, a time-of-flight detector from plastic scintillator bars with silicon photomultiplier readout and a transition radiation detector from an irregular fleece radiator and Xe/CO 2 filled proportional counting tubes. The PERDaix detector was launched with a helium balloon within the scope of the ''Balloon-Experiments for University Students'' (BEXUS) program from Kiruna, Sweden in November 2010. For a few hours PERDaix reached an altitude of 33 km and measured cosmic rays. In May 2011, the PERDaix detector was characterized during a test-beam at the PS-facilities at CERN. This work introduces methods for event

  3. CMS Hadronic Endcap Calorimeter Upgrade R&D Studies

    OpenAIRE

    Akgun, Ugur; Albayrak, Elif. A.; Onel, Yasar

    2012-01-01

    Due to an expected increase in radiation damage in LHC, we propose to replace the active material of the CMS Hadronic EndCap calorimeters with radiation hard quartz plate. Quartz is proven to be radiation hard with radiation damage tests using electron, proton, neutron and gamma beams. However, the light produced in quartz is from Cerenkov process, which yields drastically fewer photons than scintillators. To increase the light collection efficiency we pursue two separate methods: First metho...

  4. Scintillating properties of frozen new liquid scintillators

    CERN Document Server

    Britvich, G I; Golovkin, S V; Martellotti, G; Medvedkov, A M; Penso, G; Soloviev, A S; Vasilchenko, V G

    1999-01-01

    The light emission from scintillators which are liquid at room temperature was studied in the interval between $+20$~$^{\\circ}$C and $-120$~$^{\\circ}$C, where the phase transition from liquid to solid takes place. The light yield measured at $-120$~$^{\\circ}$C is about twice as much as that observed at $+20$~$^{\\circ}$C. By cooling the scintillator from $+20$~$^{\\circ}$C to $-120$~$^{\\circ}$C and then heating it from $-120$~$^{\\circ}$C to $+20$~$^{\\circ}$C, the light yield varies in steps at well defined temperatures, which are different for the cooling and heating processes. These hysteresis phenomena appear to be related to the solvent rather than to the dopant. The decay time of scintillation light was measured at $+20$~$^{\\circ}$C and $-120$~$^{\\circ}$C. Whilst at room temperature most of the light is emitted with a decay time of 6--8 ns, at $-120$~$^{\\circ}$C a slower component, with a decay time of 25--35 ns, becomes important.

  5. Detection of environmental radioactive contamination levels using a liquid-scintillation counter

    International Nuclear Information System (INIS)

    Calisto, W.; Kun, A.; Campos, E.

    1981-01-01

    A high-efficiency LS-100 C liquid scintillation counter was used to detect low levels of environmental activity. Different concentrations of primary scintillator were tested and these established the most suitable values. Work was carried out at the same time to find conditions which would ensure a low background and high efficiency. To reduce the sample volume used, various types of chelating agents were utilized: 8-hydroxyquinoline (oxine), tannic acid, cupferron, dimethylglioxime and beta-naphthol. These were tested at pH levels of 1, 6 and 11. Measurements were performed by means of the Cerenkov effect using substances with differing refraction indices - 26% sodium chloride, water, glycerine, carbon bisulphide, nitrobenzene, benzyl alcohol and toluene. Finally, work was done on comparing spectra obtained by Cerenkov radiation and by 90 Sr and 90 Y beta radiation respectively. Clearly differentiated zones were obtained, thus making it possible to distinguish one isotope from another in an equilibrium solution. (author)

  6. Hard synchrotron radiation scattering from a nonideal surface grating from multilayer X-ray mirrors

    International Nuclear Information System (INIS)

    Punegov, V.I.; Nesterets, Ya.I.; Mytnichenko, S.V.; Kovalenko, N.V.; Chernov, V.A.

    2003-01-01

    The hard synchrotron radiation scattering from a multilayer surface grating is theoretically and experimentally investigated. The numerical calculations of angular distribution of scattering intensity from X-ray mirror Ni/C are executed with use of recurrence formulae and statistical dynamical theory of diffraction. It is shown, that the essential role in formation of a diffraction pattern plays a diffuse scattering caused by structure imperfection of a multilayer grating [ru

  7. Scintillator manufacture at Fermilab

    Energy Technology Data Exchange (ETDEWEB)

    Mellott, K.; Bross, A.; Pla-Dalmau, A.

    1998-08-01

    A decade of research into plastic scintillation materials at Fermilab is reviewed. Early work with plastic optical fiber fabrication is revisited and recent experiments with large-scale commercial methods for production of bulk scintillator are discussed. Costs for various forms of scintillator are examined and new development goals including cost reduction methods and quality improvement techniques are suggested.

  8. Extruding plastic scintillator at Fermilab

    International Nuclear Information System (INIS)

    Pla-Dalmau, Anna; Bross, Alain D.; Rykalin, Viktor V.

    2003-01-01

    An understanding of the costs involved in the production of plastic scintillators and the development of a less expensive material have become necessary with the prospects of building very large plastic scintillation detectors. Several factors contribute to the high cost of plastic scintillating sheets, but the principal reason is the labor-intensive nature of the manufacturing process. In order to significantly lower the costs, the current casting procedures had to be abandoned. Since polystyrene is widely used in the consumer industry, the logical path was to investigate the extrusion of commercial-grade polystyrene pellets with dopants to yield high quality plastic scintillator. This concept was tested and high quality extruded plastic scintillator was produced. The D0 and MINOS experiments are already using extruded scintillator strips in their detectors. An extrusion line has recently been installed at Fermilab in collaboration with NICADD (Northern Illinois Center for Accelerator and Detector Development). This new facility will serve to further develop and improve extruded plastic scintillator. This paper will discuss the characteristics of extruded plastic scintillator and its raw materials, the different manufacturing techniques and the current R andD program at Fermilab

  9. Radiation hardness tests of piezoelectric actuators with fast neutrons at liquid helium temperature

    Energy Technology Data Exchange (ETDEWEB)

    Fouaidy, M.; Martinet, G.; Hammoudi, N.; Chatelet, F.; Olivier, A.; Blivet, S.; Galet, F. [CNRS-IN2P3-IPN Orsay, Orsay (France)

    2007-07-01

    Piezoelectric actuators, which are integrated into the cold tuning system and used to compensate the small mechanical deformations of the cavity wall induced by Lorentz forces due to the high electromagnetic surface field, may be located in the radiation environment during particle accelerator operation. In order to provide for a reliable operation of the accelerator, the performance and life time of piezoelectric actuators ({approx}24.000 units for ILC) should not show any significant degradation for long periods (i.e. machine life duration: {approx}20 years), even when subjected to intense radiation (i.e. gamma rays and fast neutrons). An experimental program, aimed at investigating the effect of fast neutrons radiation on the characteristics of piezoelectric actuators at liquid helium temperature (i.e. T{approx}4.2 K), was proposed for the working package WPNo.8 devoted to tuners development in the frame of CARE project. A neutrons irradiation facility, already installed at the CERI cyclotron located at Orleans (France), was upgraded and adapted for actuators irradiations tests purpose. A deuterons beam (maximum energy and beam current: 25 MeV and 35{mu}A) collides with a thin (thickness: 3 mm) beryllium target producing a high neutrons flux with low gamma dose ({approx}20%): a neutrons fluence of more than 10{sup 14} n/cm{sup 2} is achieved in {approx}20 hours of exposure. A dedicated cryostat was developed at IPN Orsay and used previously for radiation hardness test of calibrated cryogenic thermometers and pressure transducers used in LHC superconducting magnets. This cryostat could be operated either with liquid helium or liquid argon. This irradiation facility was upgraded for allowing fast turn-over of experiments and a dedicated experimental set-up was designed, fabricated, installed at CERI and successfully operated for radiation hardness tests of several piezoelectric actuators at T{approx}4.2 K. This new apparatus allows on-line automatic measurements

  10. Radiation hardness tests of piezoelectric actuators with fast neutrons at liquid helium temperature

    International Nuclear Information System (INIS)

    Fouaidy, M.; Martinet, G.; Hammoudi, N.; Chatelet, F.; Olivier, A.; Blivet, S.; Galet, F.

    2007-01-01

    Piezoelectric actuators, which are integrated into the cold tuning system and used to compensate the small mechanical deformations of the cavity wall induced by Lorentz forces due to the high electromagnetic surface field, may be located in the radiation environment during particle accelerator operation. In order to provide for a reliable operation of the accelerator, the performance and life time of piezoelectric actuators (∼24.000 units for ILC) should not show any significant degradation for long periods (i.e. machine life duration: ∼20 years), even when subjected to intense radiation (i.e. gamma rays and fast neutrons). An experimental program, aimed at investigating the effect of fast neutrons radiation on the characteristics of piezoelectric actuators at liquid helium temperature (i.e. T∼4.2 K), was proposed for the working package WPNo.8 devoted to tuners development in the frame of CARE project. A neutrons irradiation facility, already installed at the CERI cyclotron located at Orleans (France), was upgraded and adapted for actuators irradiations tests purpose. A deuterons beam (maximum energy and beam current: 25 MeV and 35μA) collides with a thin (thickness: 3 mm) beryllium target producing a high neutrons flux with low gamma dose (∼20%): a neutrons fluence of more than 10 14 n/cm 2 is achieved in ∼20 hours of exposure. A dedicated cryostat was developed at IPN Orsay and used previously for radiation hardness test of calibrated cryogenic thermometers and pressure transducers used in LHC superconducting magnets. This cryostat could be operated either with liquid helium or liquid argon. This irradiation facility was upgraded for allowing fast turn-over of experiments and a dedicated experimental set-up was designed, fabricated, installed at CERI and successfully operated for radiation hardness tests of several piezoelectric actuators at T∼4.2 K. This new apparatus allows on-line automatic measurements of actuators characteristics and the

  11. Triton X-100 as a complete liquid scintillation cocktail for counting aqueous solutions and ionic nutrient salts

    International Nuclear Information System (INIS)

    Reed, D.W.

    1984-01-01

    Triton X-100, used alone, was found to act as a complete liquid scintillation cocktail. Triton X-100 acted as a scintillator and the effect was not due to Cerenkov radiation. A variety of other commercially available surfactants also acted as scintillators, but with different levels of efficiency. Triton X-100/water combinations were suitable for counting aqueous solutions of 33 P and 86 Rb and the count rate was stable over extended periods of time. Triton X-100/toluene combinations also yielded high counting efficiencies. Triton X-100 was more sensitive to quenching than standard cocktails containing fluors. (author)

  12. Search for and selection of novel heavy scintillator crystals for calorimeter design for future high-energy colliders

    International Nuclear Information System (INIS)

    Ferrere, D.

    1993-01-01

    The discovery of some particles (Higgs, top,..) foreseen by theoretical models should be achieved at future colliders allowing to reach an energy scale of about 1 TeV. Efficient detectors must be designed to handle the very high luminosity of the LHC collider at CERN. In the intermediate mass region, M Z -2M Z , the diphoton decay mode of a Higgs boson produced inclusively or in association with W boson or a toponium gives good chance of observation. A very high resolution calorimeter with photon angle reconstruction and pion identification capability should detect a Higgs signal with high probability. So a homogeneous crystal calorimeter seems to be suitable. Because of the high luminosity and the high radiation level, a search for a new heavy scintillator has been undertaken. It must have a good radiation hardness (>0.5 MRad in a year) and a fast luminescence decay time (<30 ns). Among 50 crystals or glasses of specific chemical composition tested in transmission, luminescence, decay time, γ/neutrons radiation and light yield, cerium fluoride seems best suited for LHC. The necessity to have a good photon resolution in the intermediate Higgs mass region led us to optimise by Monte Carlo simulations the geometry of the calorimeter, the uniformisation of the light collection and crystal intercalibration parameters. (orig.)

  13. Liquid scintillation measurement. I

    International Nuclear Information System (INIS)

    Rexa, R.; Tykva, R.

    1983-01-01

    The individual components of scintillation solutions and their tasks are listed. Explained briefly is the scintillation process in a liquid scintillator. Factors are discussed which influence this process as are methods applied to supress their influence. They include: ionization quenching, quenching by dilution and concentration, chemical, colour, phase and photon quenching and single-photon events causing an undesirable backgorund. (M.D.)

  14. Characterization of the intrinsic scintillator Cs2LiCeCl6

    Energy Technology Data Exchange (ETDEWEB)

    James, R. [Savannah River Site (SRS), Aiken, SC (United States). Savannah River National Lab. (SRNL)

    2017-10-02

    In this work, we report on the scintillation properties of the intrinsic scintillator Cs2LiCeCl6 (CLCC), which is potentially useful for dual gamma-ray and neutron detection. CLCC is from the elpasolite family with a cubic structure. We grew the crystals at BNL by the vertical Bridgman growth technique. The luminescence spectrum of CLCC showed a doublet with peak maxima at 384 nm and 402 nm. The light yield of CLCC was approximately 20,000 photons/MeV, and the energy resolution was about 6% for 662-keV gamma radiation. A scintillation decay of ~81% of the total light was observed to be ~ 90 nanoseconds.

  15. Cherenkov and scintillation light separation in organic liquid scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Caravaca, J.; Descamps, F.B.; Land, B.J.; Orebi Gann, G.D. [University of California, Berkeley, CA (United States); Lawrence Berkeley National Laboratory, Berkeley, CA (United States); Yeh, M. [Brookhaven National Laboratory, Upton, NY (United States)

    2017-12-15

    The CHErenkov/Scintillation Separation experiment (CHESS) has been used to demonstrate the separation of Cherenkov and scintillation light in both linear alkylbenzene (LAB) and LAB with 2 g/L of PPO as a fluor (LAB/PPO). This is the first successful demonstration of Cherenkov light detection from the more challenging LAB/PPO cocktail and improves on previous results for LAB. A time resolution of 338 ± 12 ps FWHM results in an efficiency for identifying Cherenkov photons in LAB/PPO of 70 ± 3% and 63 ± 8% for time- and charge-based separation, respectively, with scintillation contamination of 36 ± 5% and 38 ± 4. LAB/PPO data is consistent with a rise time of τ{sub r} = 0.72 ± 0.33 ns. (orig.)

  16. Cherenkov and scintillation light separation in organic liquid scintillators

    International Nuclear Information System (INIS)

    Caravaca, J.; Descamps, F.B.; Land, B.J.; Orebi Gann, G.D.; Yeh, M.

    2017-01-01

    The CHErenkov/Scintillation Separation experiment (CHESS) has been used to demonstrate the separation of Cherenkov and scintillation light in both linear alkylbenzene (LAB) and LAB with 2 g/L of PPO as a fluor (LAB/PPO). This is the first successful demonstration of Cherenkov light detection from the more challenging LAB/PPO cocktail and improves on previous results for LAB. A time resolution of 338 ± 12 ps FWHM results in an efficiency for identifying Cherenkov photons in LAB/PPO of 70 ± 3% and 63 ± 8% for time- and charge-based separation, respectively, with scintillation contamination of 36 ± 5% and 38 ± 4. LAB/PPO data is consistent with a rise time of τ r = 0.72 ± 0.33 ns. (orig.)

  17. Mechanism of luminescent emission in BaY2F8 scintillators doped with rare earths

    International Nuclear Information System (INIS)

    Santos, Ana Carolina de Mello

    2013-01-01

    In this work samples of Barium Yttrium Fluoride (BaY 2 F 8 -BaYF) doped with different concentrations of Tb 3+ , Er 3+ , Tm 3+ e Nd 3+ ions were characterized, aiming their application in radiation detection devices using their scintillating properties. Two types of samples were produced, polycrystalline samples, obtained via solid state reaction and single crystals, obtained via the floating zone melting method. The radioluminescence measurements (RL) revealed that the samples doped with 2.0 or 3.0 % Tb showed RL intensity at 545 nm two times higher than samples of CsI:Tl, although the total scintillation yield was smaller when compared to this commercial scintillator. Thermoluminescence (TL) results showed that all doped samples exhibited the same peak at 123°C, which is characteristic of the matrix (BaY 2 F 2 ). Comparison of the areas under the TL and RL curves for Tb 3+ and Nd 3+ doped BaYF revealed that the trapping centers are competing with the scintillation of the samples. However, this was not true for the Er 3+ and Tm 3+ doped samples and it was found that the scintillation yield is independent of the trap concentration up to a certain concentration limit. From this point on, the scintillation yields started increasing as the trap concentration increased. The radiation damage defects in BaYF:Tb 3+ were studied via optical absorption (OA) and correlated with TL, showing that trapping centers are connected to the color centers. All samples exhibited different luminescence decay processes with time constants ranging from microseconds, for the fast processes, milliseconds, for the intermediate ones, and phosphorescent processes with decay time on the order of seconds. For the Nd 3+ doped samples, very fast decay process was also found with time constants of about 60ns and that is an indication that this material can be used in fast detection systems, although its scintillation efficiency was weak. The combined results obtained with these techniques together

  18. Scintillation Response of CaF2 to H and He over a Continuous Energy Range

    International Nuclear Information System (INIS)

    Zhang, Yanwen; Xiang, Xia; Weber, William J.

    2008-01-01

    Recent demands for new radiation detector materials with improved γ-ray detection performance at room temperature have prompted research efforts on both accelerated material discovery and efficient techniques that can be used to identify material properties relevant to detector performance. New material discovery has been limited due to the difficulties of large crystal growth to completely absorb γ-energies; whereas high-quality thin films or small crystals of candidate materials can be readily produced by various modern growth techniques. In this work, an ion-scintillator technique is demonstrated that can be applied to study scintillation properties of thin films and small crystals. The scintillation response of a benchmark scintillator, europium-doped calcium fluoride (CaF2:Eu), to energetic proton and helium ions is studied using the ion-scintillator approach based on a time of flight (TOF) telescope. Excellent energy resolution and fast response of the TOF telescope allow quantitative measurement of light yield, nonlinearity and energy resolution over an energy range from a few tens to a few thousands of keV

  19. Radiation-hard ASICS for sLHC optical data transmission

    International Nuclear Information System (INIS)

    Gan, K.K.

    2009-01-01

    High-speed data transmission in a high radiation environment poses an immense challenge in the detector design. We investigate the feasibility of using optical links for the silicon trackers of the ATLAS experiment for the planned upgrade of the LHC. The planned upgrade with ten times higher collision rate will produce a similar increase in the radiation. One possibility for the optical transmission is to use VCSEL arrays operating at 850 nm to transmit optical signals while using PIN arrays to convert the optical signals into electrical signals. We have designed a prototype chip containing building blocks for future SLHC optical links using a 130 nm CMOS 8RF process. The chip contains four main blocks; a VCSEL driver optimized for operation at 640 Mb/s, a VCSEL driver optimized for 3.2 Gb/s, a PIN receiver with a clock/data recovery circuit for operation at 40, 160, and 320 Mb/s, and two clock multipliers designed to operate at 640 Mb/s. The clock multiplier is designed to produce the high speed clock to serialize the data for transmission. All circuitry was designed following test results and guidelines from CERN on radiation tolerant design for the process. We have irradiated the chips with 24 GeV protons at CERN. For the VDC, the duty cycle of the output signal and the current consumption of the LVDS receiver remained constant during the irradiation. However, we observed significant decreases in the current consumption of the VCSEL driver circuit and the output drive current. This indicated that the think oxide layout used in the VCSEL driver portion of the chip might not be as radiation-hard and the circuit had been redesigned to minimize this sensitivity. For the PIN receiver, we found that the radiation produced no significant degradation, including the single event upset rate. The upset rate decreased with larger PIN current and was higher for a chip coupled to a PIN diode as expected. For the clock multipliers, we observed that the clocks of some chips

  20. Time- and wavelength-resolved luminescence evaluation of several types of scintillators using streak camera system equipped with pulsed X-ray source

    Energy Technology Data Exchange (ETDEWEB)

    Furuya, Yuki, E-mail: f.yuki@mail.tagen.tohoku.ac.j [Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577 (Japan); Yanagida, Takayuki; Fujimoto, Yutaka; Yokota, Yuui; Kamada, Kei [Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577 (Japan); Kawaguchi, Noriaki [Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577 (Japan); Research and Development Division, Tokuyama., Co. Ltd., ICR-Building, Minamiyoshinari, Aoba-ku, Sendai (Japan); Ishizu, Sumito [Research and Development Division, Tokuyama., Co. Ltd., ICR-Building, Minamiyoshinari, Aoba-ku, Sendai (Japan); Uchiyama, Koro; Mori, Kuniyoshi [Hamamatsu Photonics K.K., 325-6, Sunayama-cho, Naka-ku, Hamamatsu, Shizuoka 430-8587 (Japan); Kitano, Ken [Vacuum and Optical Instruments, 2-18-18 Shimomaruko, Ota, Tokyo 146-0092 (Japan); Nikl, Martin [Institute of Physics ASCR, Cukrovarnicka 10, Prague 6, 162-53 (Czech Republic); Yoshikawa, Akira [Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577 (Japan); NICHe, Tohoku University, 6-6-10 Aoba, Aramaki, Aoba-ku, Sendai 980-8579 (Japan)

    2011-04-01

    To design new scintillating materials, it is very important to understand detailed information about the events, which occurred during the excitation and emission processes under the ionizing radiation excitation. We developed a streak camera system equipped with picosecond pulsed X-ray source to observe time- and wavelength-resolved scintillation events. In this report, we test the performance of this new system using several types of scintillators including bulk oxide/halide crystals, transparent ceramics, plastics and powders. For all samples, the results were consistent with those reported previously. The results demonstrated that the developed system is suitable for evaluation of the scintillation properties.