WorldWideScience

Sample records for personnel neutron dosimeter

  1. The LLNL CR-39 personnel neutron dosimeter

    International Nuclear Information System (INIS)

    Hankins, D.E.; Homann, S.; Westermark, J.

    1987-01-01

    We developed a personnel neutron dosimetry system based on the electrochemical etching of CR-39 plastic at elevated temperatures. The doses obtained using this dosimeter system are more accurate than those obtained using other dosimetry systems, especially when varied neutron spectra are encountered. This CR-39 dosimetry system does not have the severe energy dependence that exists with albedo neutron dosimeters or the fading and reading problems encountered with NTA film. 3 refs., 4 figs

  2. Intercomparison of high energy neutron personnel dosimeters

    International Nuclear Information System (INIS)

    McDonald, J.C.; Akabani, G.; Loesch, R.M.

    1993-03-01

    An intercomparison of high-energy neutron personnel dosimeters was performed to evaluate the uniformity of the response characteristics of typical neutron dosimeters presently in use at US Department of Energy (DOE) accelerator facilities. It was necessary to perform an intercomparison because there are no national or international standards for high-energy neutron dosimetry. The testing that is presently under way for the Department of Energy Laboratory Accreditation Program (DOELAP) is limited to the use of neutron sources that range in energy from about 1 keV to 2 MeV. Therefore, the high-energy neutron dosimeters presently in use at DOE accelerator facilities are not being tested effectively. This intercomparison employed neutrons produced by the 9 Be(p,n) 9 B interaction at the University of Washington cyclotron, using 50-MeV protons. The resulting neutron energy spectrum extended to a maximum of approximately 50-MeV, with a mean energy of about 20-MeV. Intercomparison results for currently used dosimeters, including Nuclear Type A (NTA) film, thermoluminescent dosimeter (TLD)-albedo, and track-etch dosimeters (TEDs), indicated a wide variation in response to identical doses of high-energy neutrons. Results of this study will be discussed along with a description of plans for future work

  3. Response characteristics of selected personnel neutron dosimeters

    International Nuclear Information System (INIS)

    McDonald, J.C.; Fix, J.J.; Hadley, R.T.; Holbrook, K.L.; Yoder, R.C.; Roberson, P.L.; Endres, G.W.R.; Nichols, L.L.; Schwartz, R.B.

    1983-09-01

    Performance characteristics of selected personnel neutron dosimeters in current use at Department of Energy (DOE) facilities were determined from their evaluation of neutron dose equivalent received after irradiations with specific neutron sources at either the National Bureau of Standards (NBS) or the Pacific Northwest Laboratory (PNL). The characteristics assessed included: lower detection level, energy response, precision and accuracy. It was found that when all of the laboratories employed a common set of calibrations, the overall accuracy was approximately +-20%, which is within uncertainty expected for these dosimeters. For doses above 80 mrem, the accuracy improved to better than 10% when a common calibration was used. Individual differences found in this study may reflect differences in calibration technique rather than differences in the dose rates of actual calibration standards. Second, at dose rates above 100 mrem, the precision for the best participants was generally below +-10% which is also within expected limits for these types of dosimeters. The poorest results had a standard deviation of about +-25%. At the lowest doses, which were sometimes below the lower detection limit, the precision often approached or exceeded +-100%. Third, the lower level of detection for free field 252 Cf neutrons generally ranged between 20 and 50 mrem. Fourth, the energy dependence study provided a characterization of the response of the dosimeters to neutron energies far from the calibration energy. 11 references, 22 figures, 26 tables

  4. Personnel neutron dosimeter evaluation and upgrade program

    International Nuclear Information System (INIS)

    Fix, J.J.; Brackenbush, L.W.; McDonald, J.C.; Roberson, P.L.; Holbrook, K.L.; Endres, G.W.R.; Faust, L.G.

    1983-01-01

    Evaluation of neutron dosimeters from twelve DOE laboratories involved about 2500 dosimeter irradiations at both PNL and the National Bureau of Standards (NBS) using neutrons of several energies and doses and several irradiations for good statistical analysis. The data and their analyses will be published later. The information evaluates accuracy, precision, lower dose detection, and energy response of dosimeters

  5. Energy response study of modified CR-39 neutron personnel dosimeter

    International Nuclear Information System (INIS)

    Sathian, Deepa; Bakshi, A.K.; Datta, D.; Nair, Sreejith S.; Sathian, V.; Mishra, Jitendra; Sen, Meghnath

    2018-01-01

    Personnel neutron dosimetry is an integral part of radiation protection. No single dosimeter provides the satisfactory energy response, sensitivity, angular dependence characteristics and accuracy necessary to meet the requirement of an ideal personnel neutron dosimeter. The response of a personnel neutron dosimeter is critically dependent upon the energy distribution of the neutron field. CR-39 personnel neutron dosimeters were typically calibrated in the standard neutron field of 252 Cf and 241 Am-Be in our laboratory, although actual neutron fields may vary from the calibration neutron spectrum. Recently the badge cassette of the personnel neutron dosimeter was changed due to frequent damage of the PVC badge used earlier. This paper discusses energy response of CR-39 solid state nuclear track detector loaded in this modified badge cassette as per latest ISO recommendation

  6. Personnel neutron dosimeter for use in a plutonium processing plant

    International Nuclear Information System (INIS)

    Brunskill, R.T.; Hwang, F.S.W.

    1978-01-01

    A thermoluminesence dosimeter for personnel neutron dose measurement, which is based on the albedo principle, has been developed at Windscale works. The dosimeter has been calibrated against a 238 Pu/Be neutron source using different degrees of moderation and against a variety of neutron spectra prevailing in different areas of the Plutonium Finishing Plant. The dosimeter consists of two identical parts in which the sensitive elements are graphite discs which have thermoluminescent crystals sealed to the plane faces with a high temperature resin. The graphite discs are supported in teflon washers which fit into a body of tufnol. A circular insert of boronated polythene in each tufnol body provides a thermal neutron absorber for the sensitive element in the other half of the dosimeter. Natural lithium borate was used as the neutron sensitive phosphor and a lithium borate made from isotopes 7 Li (99.9%) and 11 B (99.2%) as the neutron insensitive materials. Neutron-sensitive lithium borate is sealed to one face of each disc and the neutron-insensitive material to the opposite face. The dosimeter is so assembled that the neutron-sensitive faces both lie in the central plane. The design is such that one neutron sensitive face responds to the incident flux of neutron only while the other responds to the albedo flux

  7. Site-specific calibration of the Hanford personnel neutron dosimeter

    International Nuclear Information System (INIS)

    Endres, A.W.; Brackenbush, L.W.; Baumgartner, W.V.; Rathbone, B.A.

    1994-10-01

    A new personnel dosimetry system, employing a standard Hanford thermoluminescent dosimeter (TLD) and a combination dosimeter with both CR-39 nuclear track and TLD-albedo elements, is being implemented at Hanford. Measurements were made in workplace environments in order to verify the accuracy of the system and establish site-specific factors to account for the differences in dosimeter response between the workplace and calibration laboratory. Neutron measurements were performed using sources at Hanford's Plutonium Finishing Plant under high-scatter conditions to calibrate the new neutron dosimeter design to site-specific neutron spectra. The dosimeter was also calibrated using bare and moderated 252 Cf sources under low-scatter conditions available in the Hanford Calibration Laboratory. Dose equivalent rates in the workplace were calculated from spectrometer measurements using tissue equivalent proportional counter (TEPC) and multisphere spectrometers. The accuracy of the spectrometers was verified by measurements on neutron sources with calibrations directly traceable to the National Institute of Standards and Technology (NIST)

  8. Personnel ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Williams, R.A.

    1975-01-01

    A dosimeter and method for use by personnel working in an area of mixed ionizing radiation fields for measuring and/or determining the effective energy of x- and gamma radiation; beta, x-, and gamma radiation dose equivalent to the surface of the body; beta, x-, and gamma radiation dose equivalent at a depth in the body; the presence of slow neutron, fast neutron dose equivalent; and orientation of the person wearing the dosimeter to the source of radiation is disclosed. Optionally integrated into this device and method are improved means for determining neutron energy spectrum and absorbed dose from fission gamma and neutron radiation resulting from accidental criticality

  9. Design characteristics of a three-component AEOI Neutriran Albedo Neutron Personnel Dosimeter

    International Nuclear Information System (INIS)

    Sohrabi, M.; Katouzi, M.

    1991-01-01

    In establishing a national personnel neutron dosimetry service in Iran, different parameters of the AEOI Neutriran Albedo Neutron Personnel Dosimeter (NANPD) have been optimized. A NANPD was designed with three dosimetry components to measure (a) direct thermal neutrons, (b) direct fast neutrons and (C) direct neutrons by the detection of the albedo neutrons reflected from the body. The dosimeter consists of one or more Lexan polycarbonate and/or CR-39 foils and two 10 B (n,α) 7 Li converters in a cadmium cover so arranged as to efficiently measure the three neutron dose components separately. The boron converter thickness, its position relative to the beam direction and its distance from the PC foil were studied and the results were incorporated into the design. The dose response of the dosimeter, its lower detection limit as well as the correction factors related to the field neutrons and albedo neutrons were also determined for a 238 Pu-Be, an 241 Am-Be and a 252 Cf sources. In this paper, the dosimeter design and its dosimetric characteristics are presented and discussed. (author)

  10. Use of an albedo neutron personnel dosimeter for X- and γ-ray monitoring

    International Nuclear Information System (INIS)

    Gorbics, S.G.; Nash, A.E.; Johnson, T.L.

    1981-01-01

    With a judicious choice of cadmium filter size and thickness, it is possible to use the information from the 7 LiF detectors used in an albedo neutron personnel dosimeter to determine an individual's X-and γ-ray exposure, thus eliminating the need for a separate dosimeter for this purpose. A filter area of 400 mm 2 and a thickness of 0.51 mm is shown to be optimum for a simple, plastic, dosimeter design using detectors held in dental-film size cards. (author)

  11. Neutron dosimeter

    International Nuclear Information System (INIS)

    Bartko, J.; Schoch, K.F. Jr.; Congedo, T.V.; Anderson, S.L. Jr.

    1989-01-01

    This patent describes a nuclear reactor. It comprises a reactor core; a thermal shield surrounding the reactor core; a pressure vessel surrounding the thermal shield; a neutron dosimeter positioned outside of the thermal shield, the neutron dosimeter comprising a layer of fissile material and a second layer made of a material having an electrical conductivity which permanently varies as a function of its cumulative ion radiation dose; and means, outside the pressure vessel and electrically connected to the layer of second material, for measuring electrical conductivity of the layer of second material

  12. Some aspects of the use of proton recoil proportional counters for fast neutron personnel dosimeters

    International Nuclear Information System (INIS)

    Yule, T.J.; Bennett, E.F.

    1984-01-01

    Gas-filled proton recoil proportional counters have been used extensively for the measurement of neutron spectra in degraded fission-spectrum environments. Some considerations relating to the use of these counters for personnel dosimetry are here described. High sensitivity and good accuracy in the determination of dose-equivalent can be obtained if relatively high pressure hydrogen-filled proportional counters are used as the active element in a dosimeter system

  13. Hanford personnel dosimeter supporting studies FY-1981

    International Nuclear Information System (INIS)

    1982-08-01

    This report examined specific functional components of the routine external personnel dosimeter program at Hanford. Components studied included: dosimeter readout; dosimeter calibration; dosimeter field response; dose calibration algorithm; dosimeter design; and TLD chip acceptance procedures. Additional information is also presented regarding the dosimeter response to light- and medium-filtered x-rays, high energy photons and neutrons. This study was conducted to clarify certain data obtained during the FY-1980 studies

  14. 1987 Neutron and gamma personnel dosimeter intercomparison study using a D2O-moderated 252Cf source

    International Nuclear Information System (INIS)

    Swaja, R.E.; West, L.E.; Sims, C.S.; Welty, T.J.

    1989-05-01

    The thirteenth Personnel Dosimetry Intercomparison Study (i.e., PDIS 13) was conducted during April 1987 as a joint effort by Oak Ridge National Laboratory's (ORNL) Dosimetry Applications Research Group and the Southwest Radiation Calibration Center at the University of Arkansas. A total of 48 organizations (34 from the US and 14 from abroad) participated in PDIS 13. Participants submitted a total of 1,113 neutron and gamma dosimeters for this mixed field study. The dosimeters were transferred by mail and were handled by experimental personnel at ORNL and the University of Arkansas. The type of neutron dosimeter and the percentage of participants submitting that type are as follows: TLD-albedo (49%), direct interaction TLD (31%), CR-39 (17%), film (3%). The type of gamma dosimeter and the percentage of participants submitting that type are as follows: Li 2 B 4 O 7 , alone or in combination with CaSO 4 , (69%), 7 LiF (28%), natural LiF (3%). Radiation exposures in PDIS 13 were limited to 0.5 and 1.5 mSv from 252 Cf moderated by 15-cm of D 2 O. Traditional exposures using the Health Physics Research Reactor (HPRR) were not possible due to the fact that all reactors at ORNL, including the HPRR, were shutdown by order of the Department of Energy at the time the intercomparison was performed. Planned exposures using a 238 PuBe source were negated by a faulty timing mechanism. Based on accuracy and precision, direct interaction TLD dosimeters exhibited the best performance in PDIS 13 neutron measurements. They were followed, in order of best performance, by CR-39, TLD albedo, and film. The Li 2 B 4 O 7 type TLD dosimeters exhibited the best performance in PDIS 13 gamma measurements. They were followed by natural LiF, 7 LiF, and film. 12 refs., 1 fig., 5 tabs

  15. Response of combined albedo-track neutron personnel dosimeters behind IHEP proton synchrotron shielding

    International Nuclear Information System (INIS)

    Sannikov, A.V.; Korshunova, E.P.

    1989-01-01

    The method of readings interpretation of combined albedo-track neutron personnel dosemeters based on calculationsl analysis of the detector responses in various neutron spectra is described. The measurements of dose equivalent responses have been performed in various points behind IHEP proton synchrotron shielding. It is shown that CDs with fission track detectors have a small dose equivalent response dispersion behind IHEP proton synchrotron shielding, that shows the promise of their using for neutron personnel monitoring, that shows the promise of their using for neutron personnel monitoring at high energy accelerators. 16 refs.; 7 figs.; 3 tabs

  16. Neutron dosimeter utilizing CR-39

    International Nuclear Information System (INIS)

    Souza, H.V.C. de.

    1991-05-01

    A personnel neutron dosimeter has been developed with discretization in a wide range of energies of real interest, utilizing the CR-39 polymer, to detect recoil protons in the fast range, and alpha particles in the thermal and epithermal ranges, with possibility to be disposed in the IRD/CNEN's conventional film badge suport. They are presented, abstractly, the difficulties and importance of the neutron dosimetry, beyond the general objectives that motivated this work execution. The details of the materials utilized in the dosimeter confection, and the experimental methodology employed to obtain the performance curves are presented. The results about linearity response of the dosimeter with respect to equivalent dose, in a wide range of doses, and about the verified angular dependence are analysed. (author)

  17. New Neutron Dosimeter

    CERN Multimedia

    2001-01-01

    CERN has been operating an Individual Dosimetry Service for neutrons for about 35 years. The service was based on nuclear emulsions in the form of film packages which were developed and scanned in the Service. In 1999, the supplier of theses packages informed CERN that they will discontinue production of this material. TIS-RP decided to look for an external service provider for individual neutron dosimetry. After an extensive market survey and an invitation for tender, a supplier that met the stringent technical requirements set up by CERN's host states for personal dosimeters was identified. The new dosimeter is based on a track-etching technique. Neutrons have the capability of damaging plastic material. The microscopic damage centres are revealed by etching them in a strong acid. The resulting etch pits can be automatically counted and their density is proportional to dose equivalent from neutrons. On the technical side, the new dosimeter provides an improved independence of its response from energy and th...

  18. Guidelines for the calibration of personnel dosimeters

    International Nuclear Information System (INIS)

    Roberson, P.L.; Holbrook, K.L.

    1984-01-01

    This guide describes minimum acceptable performance levels for personnel dosimetry systems used at Department of Energy (DOE) facilities. The goal is to improve both the quality of radiological calibrations and the methods of comparing reported occupational doses between DOE facilities. Reference calibration techniques are defined. A standard for evaluation of personnel dosimetry systems and recommended design parameters for personnel dosimeters are also included. Approximate intervals for the radiation energies for which these guidelines are appropriate are 15 keV to 2 MeV for photons; above 0.3 MeV for beta particles; and 1 keV to 2 MeV for neutrons. An analysis of ANSI N13.11 was completed using performance evaluations of selected personnel dosimetry systems in use at DOE facilities. The results of this analysis are incorporated in the guidelines

  19. An Emergency Dosimeter for Neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Braun, J; Nilsson, R

    1960-05-15

    A neutron dosimeter suitable for single emergency exposures is described. The dosimeter is furnished with detectors for thermal, epi-thermal and fast neutrons. This means that three of the constants by which the spectrum of the incident neutron flux is approximated, can be determined. The dose calculated from these approximated spectra is compared to the dose from spectra obtained in different standard spectra of types which may be expected in a radiation accident.

  20. Digital neutron dosimeter

    International Nuclear Information System (INIS)

    Ramondetta, P.W.; Groeber, E.O.Jr.

    1978-01-01

    Design features for a portable battery-operated neutron dosimeter are described. The system employs a 50-mil PIN detector diode, whose forward voltage increases with exposure to fast neutrons. Because this change is permanent and cumulative, the system is able to integrate small doses (from 0 to 1000 rad) over long periods of time. The system is temperature compensated over its operating range of -40 C to +52C. Display accuracies of +-20 rad for readings below 100 rad and +-20% for readings above 100 rad are maintained throughout the range. Temperature correction is performed digitally after an initial analog-to-digital conversion of both the forward diode voltage and the ambient temperature. System flexibility is promoted with the use of a replaceable ROM for the final voltage-to-dose conversion table. This digital approach to temperature compensation, combined with the extensive use of CMOS circuitry, suggests the use of custom large-scale integration as a means of further reducing system weight and size. This possibility, as well as others, is discussed as a means of reducing system size. Test and evaluation results are also included. (author)

  1. LOW-COST PERSONNEL DOSIMETER.

    Science.gov (United States)

    specification was achieved by simplifying and improving the basic Bendix dosimeter design, using plastics for component parts, minimizing direct labor, and making the instrument suitable for automated processing and assembly. (Author)

  2. Neutron personnel dosimetry

    International Nuclear Information System (INIS)

    Griffith, R.V.

    1982-01-01

    The measurement of neutron exposures to personnel is an issue that has received increased attention in the last few years. It is important to consider key aspects of the whole dosimetry system when developing dose estimates. This begins with selection of proper dosimeters and survey instruments, and extends through the calibration methods. One must match the spectral response and sensitivity of the dosimeter to the spectral characteristics of the neutron fields. Threshold detectors that are insensitive to large fractions of neutrons in the lower energy portion of reactor spectra should be avoided. Use of two or more detectors with responses that complement each other will improve measurement quality. It is important to understand the spectral response of survey instruments, so that spectra which result in significant overresponse do not lead to overestimation of dose. Calibration sources that do not match operational field spectra can contribute to highly erroneous results. In those situations, in-field calibration techniques should be employed. Although some detection developments have been made in recent years, a lot can be done with existing technology until fully satisfactory, long term solutions are obtained

  3. VALIDATION OF HANFORD PERSONNEL AND EXTREMITY DOSIMETERS IN PLUTONIUM ENVIRONMENTS

    Energy Technology Data Exchange (ETDEWEB)

    Scherpelz, Robert I.; Fix, John J.; Rathbone, Bruce A.

    2000-02-10

    A study was performed in the Plutonium Finishing Plant to assess the performance of Hanford personnel neutron dosimetry. The study was assessed whole body dosimetry and extremity dosimetry performance. For both parts of the study, the TEPC was used as the principle instrument for characterizing workplace neutron fields. In the whole body study, 12.7-cm-diameter TEPCs were used in ten different locations in the facility. TLD and TED personnel dosimeters were exposed on a water-filled phantom to enable a comparison of TEPC and dosimeter response. In the extremity study, 1.27-cm-diameter TEPCs were exposed inside the fingers of a gloveboxe glove. Extremity dosimeters were wrapped around the TEPCs. The glove was then exposed to six different cans of plutonium, simulating the exposure that a worker's fingers would receive in a glovebox. The comparison of TEPC-measured neutron dose equivalent to TLD-measured gamma dose equivalent provided neutron-to-gamma ratios that can be used to estimate the neutron dose equivalent received by a worker's finger based on the gamma readings of an extremity dosimeter. The study also utilized a Snoopy and detectors based on bubble technology for assessing neutron exposures, providing a comparison of the effectiveness of these instruments for workplace monitoring. The study concludes that the TLD component of the HCND performs adequately overall, with a positive bias of 30%, but exhibits excessive variability in individual results due to instabilities in the algorithm. The TED response was less variable but only 20% of the TEPC reference dose on average because of the low neutron energies involved. The neutron response of the HSD was more variable than the TLD component of the HCND and biased high by a factor of 8 overall due to its calibration to unmoderated 252Cf. The study recommends further work to correct instabilities in the HCND algorithm and to explore the potential shown by the bubble-based dosimeters.

  4. Systematic approach to personnel neutron monitoring

    International Nuclear Information System (INIS)

    Griffith, R.V.; Hankins, D.E.

    1980-01-01

    NTA film and albedo detectors represent the major portion of personnel dosimeters now used for occupational neutron monitoring. However, recent attention to the spectral response of these systems has demonstrated the need for detectors that have a better match to the fields being monitored. Recent developments in direct recoil track etch dosimeters present some intriguing alternatives, and careful use of 237 Np fission fragment detectors offers the advantage of a good dose equivalent spectral match. Work continues on a number of other new detector mechanisms, but problems with sensitivity, energy response, gamma interference, etc., continue to prevent development of most mechanisms into viable personnel dosimeters. Current dosimeter limitations make a systematic approach to personnel neutron monitoring particularly important. Techniques have been developed and tested, using available portable survey instruments, that significantly improve the quality of dosimeter interpretation. Even simple spectrometry can be done with modest effort, significantly improving the health physicists ability to provide accurate neutron monitoring

  5. Automating the personnel dosimeter monitoring program

    International Nuclear Information System (INIS)

    Compston, M.W.

    1982-12-01

    The personnel dosimetry monitoring program at the Portsmouth uranium enrichment facility has been improved by using thermoluminescent dosimetry to monitor for ionizing radiation exposure, and by automating most of the operations and all of the associated information handling. A thermoluminescent dosimeter (TLD) card, worn by personnel inside security badges, stores the energy of ionizing radiation. The dosimeters are changed-out periodically and are loaded 150 cards at a time into an automated reader-processor. The resulting data is recorded and filed into a useful form by computer programming developed for this purpose

  6. High energy neutron dosimeter

    International Nuclear Information System (INIS)

    Rai, K.S.F.

    1994-01-01

    A device for measuring dose equivalents in neutron radiation fields is described. The device includes nested symmetrical hemispheres (forming spheres) of different neutron moderating materials that allow the measurement of dose equivalents from 0.025 eV to past 1 GeV. The layers of moderating material surround a spherical neutron counter. The neutron counter is connected by an electrical cable to an electrical sensing means which interprets the signal from the neutron counter in the center of the moderating spheres. The spherical shape of the device allows for accurate measurement of dose equivalents regardless of its positioning. 2 figures

  7. US progress on the development of CR-39 based neutron dosimeters

    International Nuclear Information System (INIS)

    Hadlock, D.E.

    1987-06-01

    Historically at US nuclear facilities, two types of personnel neutron dosimeters have been in routine use: nuclear track emulsion-Type A (NTA) film and thermoluminescent dosimeter (TLD)-albedo. Both of these dosimeters have energy-dependent responses. Therefore, the neutron energy spectra must be known, to interpret the dosimeter results properly. A new state-of-the-art dosimetry system has been developed within the US Department of Energy (US DOE) Personnel Neutron Dosimeter Evaluation and Upgrade Program. This system is called the combination thermoluminescent dosimeter/track etch dosimeter (TLD/TED). This paper briefly describes US DOE research currently being conducted to further enhance the TED portion of the combination TLD/TED system. The research areas involved include dose sensitivity, neutron energy range, specialized radiators, self-developing dosimeters, and neutron spectrometry. 1 fig., 1 tab

  8. A pocket type thermoluminescent personnel dosimeter

    International Nuclear Information System (INIS)

    Vora, K.G.; Nagpal, J.S.; Pendurkar, H.K.; Gangadharan, P.

    1979-01-01

    A pocket type thermoluminescent personnel dosemeter using CaSO 4 : Dy phosphor is described. Two glass capillaries containing the phosphor are fitted into a plastic tube and covered by a cylindrical filter. The combination is fitted into an ink barrel of a fountain pen. The response of this Dy glass dosimeter was studied for various incident photon energies. A uniform response over the energy range from 33 keV to 1250 keV is achieved. (A.K.)

  9. Neutron personnel dosimetry

    International Nuclear Information System (INIS)

    Griffith, R.V.

    1981-01-01

    The current state-of-the-art in neutron personnel dosimetry is reviewed. Topics covered include dosimetry needs and alternatives, current dosimetry approaches, personnel monitoring devices, calibration strategies, and future developments

  10. A systematic approach to personnel neutron monitoring

    International Nuclear Information System (INIS)

    Griffith, R.V.; Hankins, D.E.

    1980-01-01

    In selection, calibration and interpretation of personnel neutron dosimeters used in radiation protection, adequate attention is often not given to matching the characteristics of the dosimeter with the quality of the neutron field. A particular concern is the use of albedo detectors which have little energy response similarity to the neutron dose equivalent conversion curve. At the Lawrence Livermore Laboratory we have developed a system for dosimeter calibration and neutron field characterization using Bonner spheres and remmeters. Rapid surveys of the work area with detectors in 3-in and 9-in polyethylene spheres establish a qualitative estimate of spectral variation found in the facility. We also use this data to determine the appropriate albedo dosimeter calibration factors. At several locations representing the spectral range, multisphere spectra measure-ments are made and the spectrum weighted dose equivalent rates calculated. These rates are compared with survey instrument results to establish correction factors for the relative over- or under-response expected from these instruments, particularly in highly moderated neutron fields where remmeters overrespond. We also use the spectral information to determine the appropriateness of dosimeters considered for future use. This technique has been applied at power reactors to provide information valuable to selection of proper personnel dosimeters. We find that the spectral range is sufficiently narrow that albedo detectors can be used with confidence. On the other hand, most of the dose occurs at energies below the effective threshold NTA film. (author)

  11. Study of a plastic detector as a neutron dosimeter

    International Nuclear Information System (INIS)

    Cuauhtecatl Hernandes, V.

    1982-01-01

    Studies were carried out through nuclear reactions (n, p); (n, heavy ion), and (n,α) of the dosimetric properties of CR 39 commercial brand polymers. A system was devised for calculating the number of neutron induced nuclear reactions and geometric absolute efficiency factors. Feasibility of the utilization of CR 39 in monitoring and in personnel dosimeters is also discussed. (author)

  12. Personal neutron dosimeter using solid-state track detector

    International Nuclear Information System (INIS)

    Mettripan, S.

    1980-01-01

    A cellulose nitrate film coated on both sides with lithium tetraborate was used as a neutron dosimeter for surveillance of personnel exposed to thermal and epithermal neutron. It was found that the optimum etching conditions used were 10% solution of sodium hydroxide, 60 degrees C and 20 minutes etching time and the alpha track densities from the (n,α) reaction on the films were proportional to thermal and epithermal neutron fluxes. The response of the film was found to be 1.068 x 10 -3 tracks per thermal neutron and 3.438 x 10 -4 tracks per epithermal neutron

  13. Use of wrist albedo neutron dosimeters

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1983-01-01

    We are developing a wrist dosimeter that can be used to measure the exposure at the wrist to x-rays, gamma rays, beta-particles, thermal neutrons and fast neutrons. It consists of a modified Hankins Type albedo neutron dosimeter and also contains three pieces of CR-39 plastic. ABS plastic in the form of an elongated hemisphere provides the beta and low energy x-ray shielding necessary to meet the requirement of depth dose measurements at 1 cm. The dosimeter has a beta window located in the side of the hemisphere oriented towards an object being held in the hands. A TLD 600 is positioned under the 1 cm thick ABS plastic and is used to measure the thermal neutron dose. At present we are using Velcro straps to hold the dosimeter on the inside of the wrist. 9 figures

  14. Fast-neutron solid-state dosimeter

    International Nuclear Information System (INIS)

    Kecker, K.H.; Haywood, F.F.; Perdue, P.T.; Thorngate, J.H.

    1975-01-01

    This patent relates to an improved fast-neutron solid-state dosimeter that does not require separation of materials before it can be read out, that utilizes materials that do not melt or otherwise degrade at about 300 0 C readout temperature, that provides a more efficient dosimeter, and that can be reused. The dosimeters are fabricated by intimately mixing a TL material, such as CaSO 4 :Dy, with a powdered polyphenyl, such as p-sexiphenyl, and hot-pressing the mixture to form pellets, followed by out-gassing in a vacuum furnace at 150 0 C prior to first use dosimeters

  15. Personnel neutron dosimetry at Department of Energy facilities

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Endres, G.W.R.; Selby, J.M.; Vallario, E.J.

    1980-08-01

    This study assesses the state of personnel neutron dosimetry at DOE facilities. A survey of the personnel dosimetry systems in use at major DOE facilities was conducted, a literature search was made to determine recent advances in neutron dosimetry, and several dosimetry experts were interviewed. It was concluded that personnel neutron dosimeters do not meet current needs and that serious problems exist now and will increase in the future if neutron quality factors are increased and/or dose limits are lowered

  16. Personnel neutron dosimetry

    International Nuclear Information System (INIS)

    Hankins, D.

    1982-04-01

    This edited transcript of a presentation on personnel neutron discusses the accuracy of present dosimetry practices, requirements, calibration, dosemeter types, quality factors, operational problems, and dosimetry for a criticality accident. 32 figs

  17. Personnel neutron dose assessment upgrade: Volume 1, Personnel neutron dosimetry assessment: [Final report

    International Nuclear Information System (INIS)

    Hadlock, D.E.; Brackenbush, L.W.; Griffith, R.V.; Hankins, D.E.; Parkhurst, M.A.; Stroud, C.M.; Faust, L.G.; Vallario, E.J.

    1988-07-01

    This report provides guidance on the characteristics, use, and calibration criteria for personnel neutron dosimeters. The report is applicable for neutrons with energies ranging from thermal to less than 20 MeV. Background for general neutron dosimetry requirements is provided, as is relevant federal regulations and other standards. The characteristics of personnel neutron dosimeters are discussed, with particular attention paid to passive neutron dosimetry systems. Two of the systems discussed are used at DOE and DOE-contractor facilities (nuclear track emulsion and thermoluminescent-albedo) and another (the combination TLD/TED) was recently developed. Topics discussed in the field applications of these dosimeters include their theory of operation, their processing, readout, and interpretation, and their advantages and disadvantages for field use. The procedures required for occupational neutron dosimetry are discussed, including radiation monitoring and the wearing of dosimeters, their exchange periods, dose equivalent evaluations, and the documenting of neutron exposures. The coverage of dosimeter testing, maintenance, and calibration includes guidance on the selection of calibration sources, the effects of irradiation geometries, lower limits of detectability, fading, frequency of calibration, spectrometry, and quality control. 49 refs., 6 figs., 8 tabs

  18. Intercomparison measurements with albedo neutron dosimeters

    International Nuclear Information System (INIS)

    Alberts, W.G.; Kluge, H.

    1994-01-01

    Since the introduction of the albedo dosimeter as the official personal neutron dosimeter the dosimetry services concerned have participated in intercomparison measurements at the PTB. Their albedo dosimeters were irradiated in reference fields produced by unmoderated and D 2 O-moderated 252 Cf neutron sources in the standard irradiation facility of the PTB. Six fields with fluences different in energy and angle distribution could be realised in order to determine the response of the albedo dosimeter. The dose equivalent values evaluated by the services were compared with the reference values of the PTB for the directional dose equivalent H'(10). The results turned out to be essentially dependent on the evaluation method and the choice of the calibration factors. (orig.) [de

  19. Hanford beta-gamma personnel dosimeter prototypes and evaluation

    International Nuclear Information System (INIS)

    Fix, J.J.; Holbrook, K.L.; Soldat, K.L.

    1983-04-01

    Upgraded and modified Hanford dosimeter prototypes were evaluated for possible use at Hanford as a primary beta-gamma dosimeter. All prototypes were compatible with the current dosimeter card and holder design, as well as processing with the automated Hanford readers. Shallow- and deep-dose response was determined for selected prototypes using several beta sources, K-fluorescent x rays and filtered x-ray techniques. All prototypes included a neutron sensitive chip. A progressive evaluation of the performance of each of the upgrades to the current dosimeter is described. In general, the performance of the current dosimeter can be upgraded using individual chip sensitivity factors to improve precision and an improved algorithm to minimize bias. The performance of this dosimeter would be adequate to pass all categories of the ANSI N13.11 performance criteria for dosimeter procesors, provided calibration techniques compatible with irradiations adopted in the standard were conducted. The existing neutron capability of the dosimeter could be retained. Better dosimeter performance to beta-gamma radiation can be achieved by modifying the Hanford dosimeter so that four of the five chip positions are devoted to calculating these doses instead of the currently used two chip positions. A neutron sensitive chip was used in the 5th chip position, but all modified dosimeter prototypes would be incapable of discriminating between thermal and epithermal neutrons. An improved low energy beta response can be achieved for the current dosimeter and all prototypes considered by eliminating the security credential. Further improvement can be obtained by incorporating the 15-mil thick TLD-700 chips

  20. The active personnel dosimeter---APFEL enterprises superheated drop detector

    International Nuclear Information System (INIS)

    Ipe, N.E.; Donahue, R.J.; Busick, D.D.

    1991-03-01

    The Active Personnel Dosimeter (APD) provides a digital readout of events caused by neutrons interacting with superheated liquid droplets. The droplets are suspended in a gel held in a replaceable cartridge. Upon neutron interaction, the superheated droplet vaporizes, forming a bubble. The sound produced in this process is recorded by transducers that sense the accompanying pressure pulse. The APD electronically discriminates against spurious noise and vibration. Studies with the production prototype APDs indicate that the detector response is linear up to about 0.40 mSv, with large variations sometimes from predicted values and between cartridges at higher dose equivalents. The response to standard neutron sources (bare 252 Cf, PuBe, PuB, PuF, PuLi) is reported and compared with the expected response. Unirradiated cartridges self-nucleate when heated to temperatures of 46 degrees C. The APD is insensitive to low-energy photons but responds to high-energy photons and electrons. 8 refs., 2 figs., 3 tabs

  1. Personnel neutron dosimetry using electrochemically etched CR-39 foils

    International Nuclear Information System (INIS)

    Hankins, D.E.; Homann, S.; Westermark, J.

    1986-01-01

    A personnel neutron dosimetry system has been developed based on the electrochemical etching of CR-39 plastic at elevated temperatures. The doses obtained using this dosimeter system are more accurate than those obtained using other dosimetry systems, especially when varied neutron spectra are encountered. This Cr-39 dosimetry system does not have the severe energy dependence that exists with albedo neutron dosimeters or the fading and reading problems encountered with NTA film. The dosimetry system employs an electrochemical etch procedure that be used to process large numbers of Cr-39 dosimeters. The etch procedure is suitable for operations where the number of personnel requires that many CR-39 dosimeters be processed. Experience shows that one full-time technician can etch and evaluate 2000 foils per month. The energy response to neutrons is fairly flat from about 80 keV to 3.5 MeV, but drops by about a factor of three in the 13 to 16 MeV range. The sensitivity of the dosimetry system is about 7 tracks/cm 2 /mrem, with a background equivalent to about 8 mrem for new CR-39 foils. The limit of sensitivity is approximately 10 mrem. The dosimeter has a significant variation in directional dependence, dropping to about 20% at 90 0 . This dosimeter has been used for personnel neutron dosimetry at the Lawrence Livermore National Laboratory for more tha 18 months. 6 refs., 23 figs., 2 tabs

  2. Evaluation of the neutron dose received by personnel at the LLNL

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1982-01-01

    This report was prepared to document the techniques being used to evaluate the neutron exposures received by personnel at the LLNL. Two types of evaluations are discussed covering the use of the routine personnel dosimeter and of the albedo neutron dosimeter. Included in the report are field survey results which were used to determine the calibration factors being applied to the dosimeter readings. Calibration procedures are discussed and recommendations are made on calibration and evaluation procedures

  3. Silicon Diode Dosimeter for Fast Neutrons

    International Nuclear Information System (INIS)

    Svansson, L.; Widell, C.O.; Swedberg, P.; Wik, M.

    1968-11-01

    The change of the current-voltage characteristics of a small silicon diode is used as a measure of fast neutron dose in the Fast Neutron Dosimeter 5422. This change is permanent and therefore it is possible to integrate doses over a long period of time. Doses from some rad up to 1000 rad can be measured and the information stored is not destroyed during readout. Considerable research work in this field has previously been carried out by the Swedish Institute for National Defence in collaboration with the Institute of Semiconductor Research Stockholm. The present investigation has been made in order to establish the possibilities of the dosimeter for practical applications and to study the variations of important parameters as a function of the production process. In particular the following parameters have been studied: - dose sensitivity, - energy dependence; - fading effect; - temperature influence; - maximum measurable dose. In general one might conclude that the dosimeter 5422 well fulfills requirements usually specified for a dosimeter for field service. Temperature influence and fading effect are of little practical importance within the recommended range of measurement

  4. Silicon Diode Dosimeter for Fast Neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Svansson, L; Widell, C O; Swedberg, P [The Inst. of Semiconductor Researc h, Stockholm (Sweden); Wik, M [The Swedish Institute for National Defence, Sun dbyberg (Sweden)

    1968-11-15

    The change of the current-voltage characteristics of a small silicon diode is used as a measure of fast neutron dose in the Fast Neutron Dosimeter 5422. This change is permanent and therefore it is possible to integrate doses over a long period of time. Doses from some rad up to 1000 rad can be measured and the information stored is not destroyed during readout. Considerable research work in this field has previously been carried out by the Swedish Institute for National Defence in collaboration with the Institute of Semiconductor Research Stockholm. The present investigation has been made in order to establish the possibilities of the dosimeter for practical applications and to study the variations of important parameters as a function of the production process. In particular the following parameters have been studied: - dose sensitivity, - energy dependence; - fading effect; - temperature influence; - maximum measurable dose. In general one might conclude that the dosimeter 5422 well fulfills requirements usually specified for a dosimeter for field service. Temperature influence and fading effect are of little practical importance within the recommended range of measurement.

  5. Eleventh DOE workshop on personnel neutron dosimetry

    International Nuclear Information System (INIS)

    1991-01-01

    Since its formation, the Office of Health (EH-40) has stressed the importance of the exchange of information related to and improvements in neutron dosimetry. This Workshop was the eleventh in the series sponsored by the Department of Energy (DOE). It provided a forum for operational personnel at DOE facilities to discuss current issues related to neutron dosimetry and for leading investigators in the field to discuss promising approaches for future research. A total of 26 papers were presented including the keynote address by Dr. Warren K. Sinclair, who spoke on, ''The 1990 Recommendations of the ICRP and their Biological Background.'' The first several papers discussed difficulties in measuring neutrons of different energies and ways of compensating or deriving correction factors at individual facilities. Presentations were also given by the US Navy and Air Force. Current research in neutron dosimeter development was the subject of the largest number of papers. These included a number on the development of neutron spectrometers

  6. The neutron response of a 7 LiF thermoluminescent dosimeter incorporated in the UKAEA criticality dosimeter

    International Nuclear Information System (INIS)

    Eid, A.M.; Delafield, H.J.

    1976-04-01

    There are practical advantages in incorporating a 7 LiF thermoluminescent dosimeter (TLD) for the measurement of γ-ray dose, into the personnel criticality dosimeter. This paper investigated the corrections necessary for the inherent direct response of the TLD neutrons, and its enhanced indirect response from prompt γ-rays resulting from neutron interactions with the metallic foils contained in the criticality dosimeter. The response of the TLD to fast fission neutrons was measured to be 0.02 γ rad/n rad. The indirect response of the TLD to thermal neutrons was measured to be 4.8 x 10 -10 rad n -1 cm 2 for dosimeters exposed in free air, and 7 x 10 -10 rad n -1 cm 2 for dosimeters worn on the body respectively. Application of these correction factors to TLD measurements made at International Dosimetry Intercomparisons (sponsored by the I.A.E.A.) gave improved agreement with the values given by other participants. (author)

  7. Role of gel dosimeters in boron neutron capture therapy

    International Nuclear Information System (INIS)

    Khajeali, Azim; Farajollahi, Ali Reza; Khodadadi, Roghayeh; Kasesaz, Yaser; Khalili, Assef

    2015-01-01

    Gel dosimeters have acquired a unique status in radiotherapy, especially with the advent of the new techniques in which there is a need for three-dimensional dose measurement with high spatial resolution. One of the techniques in which the use of gel dosimeters has drawn the attention of the researchers is the boron neutron capture therapy. Exploring the history of gel dosimeters, this paper sets out to study their role in the boron neutron capture therapy dosimetric process. - Highlights: • Gel dosimeters have been investigated. • Conventional dosimetric proses of BNCT has been investigated. • Role of gel dosimeters in BNCT has been investigated

  8. Study of an individual neutron dosimeter

    International Nuclear Information System (INIS)

    Debeauvais, M.; Tripier, J.

    1976-01-01

    A dosimeter using Kodak LR 115 cellulose nitrate as detecting material was designed. It serves to determine 3 neutron energy ranges. The 6 Li(n,α)t reaction is used for the thermal region, the sensitivity being 0.2mrads to 1 rad for neutron energies between thermal and 0.05eV. The same reaction defines the 0.05eV to 1000eV energy range but the detection system is placed inside a cadmium screen; the sensitivity is 0.2 to 500rads. Finally above 1MeV the neutron reactions used are those on the detector components themselves, i.e. elastic collisions and (nα) reactions on carbon, nitrogen and oxygen nuclei. Detection is possible between 0.7 and 700 rads [fr

  9. Fast neutron dosimeter with wide base silicon diode

    International Nuclear Information System (INIS)

    Ma Lu

    1986-01-01

    This paper briefly introduces a wide base silicon diode fast neutron dosimeter with wide measuring range and good energy response to fast neutron. It is suitable to be used to detect fast neutrons in the mixed field of γ-ray, thermal neutrons and fast neutrons

  10. Advances in the development of Cr-39 based neutron dosimeters

    International Nuclear Information System (INIS)

    Hadlock, D.E.; Parkhurst, M.A.

    1987-12-01

    A combination thermoluminescent dosimeter (TLD) and track etch dosimeter (TED), which can be used for detecting neutrons over a wide energy range, has been developed through recent research in passive neutron dosimetery. This dosimeter uses Li-600 TLDs to detect thermal and low energy neutrons reflected from the body, and the TED polymer of CR-39, to detect fast neutrons from proton recoil interactions with the polyethylene radiator or with CR-39 itself. Some form of the combination dosimeter is currently in use at several US Department of Energy (DOE) facilities, and its use is expected to expand over the next year to include all DOE facilities where significant neutron exposures may occur. The extensive research conducted on the TED component over the past six years has continually focused on material improvements, reduction in processing time and dosimeter handling, and ease of sample readout with the goal of automating the process as much as possible. 1 fig

  11. Personnel Neutron Monitoring at AB Atomenergi

    Energy Technology Data Exchange (ETDEWEB)

    Hagsgaard, S; Widell, C O

    1964-02-15

    The routine personnel monitoring of fast neutrons is carried out by the counting of tracks in a nuclear emulsion. The tracks are counted in a microscope on a projection screen. This is a very tedious job and is only done on irradiated films which are counted over 6 mm{sup 2}. The irradiated films are selected according to the recorded dose on the gamma film. It is often difficult to tell how much the visible tracks have faded during a two-weeks period. Fortunately the fading does not often exceed 20 % for this period. If the dosimeter has been gamma-irradiated, it may be difficult to recognize the proton tracks. If the film is stored for some time before being developed, this gamma fog will to some extent fade away. For large neutron doses a foil activation dosimeter is used. This dosimeter consists of a cadmium-shielded phosphorus foil, a cadmium shielded gold foil and an unshielded gold foil. The phosphorus foil has to be counted shortly after exposure.

  12. Personnel Neutron Monitoring at AB Atomenergi

    International Nuclear Information System (INIS)

    Hagsgaard, S.; Widell, C.O.

    1964-02-01

    The routine personnel monitoring of fast neutrons is carried out by the counting of tracks in a nuclear emulsion. The tracks are counted in a microscope on a projection screen. This is a very tedious job and is only done on irradiated films which are counted over 6 mm 2 . The irradiated films are selected according to the recorded dose on the gamma film. It is often difficult to tell how much the visible tracks have faded during a two-weeks period. Fortunately the fading does not often exceed 20 % for this period. If the dosimeter has been gamma-irradiated, it may be difficult to recognize the proton tracks. If the film is stored for some time before being developed, this gamma fog will to some extent fade away. For large neutron doses a foil activation dosimeter is used. This dosimeter consists of a cadmium-shielded phosphorus foil, a cadmium shielded gold foil and an unshielded gold foil. The phosphorus foil has to be counted shortly after exposure

  13. Solid state neutron dosimeter for space applications. Final Report

    International Nuclear Information System (INIS)

    Entine, G.; Nagargar, V.; Sharif, D.

    1990-08-01

    Personnel engaged in space flight are exposed to significant flux of high energy neutrons arising from both primary and secondary sources of ionizing radiation. Presently, there exist no compact neutron sensor capable of being integrated in a flight instrument to provide real time measurement of this radiation flux. A proposal was made to construct such an instrument using special PIN silicon diode which has the property of being insensitive to the other forms of ionizing radiation. Studies were performed to determine the design and construction of a better reading system to allow the PIN diode to be read with high precision. The physics of the device was studied, especially with respect to those factors which affect the sensitivity and reproducibility of the neutron response. This information was then used to develop methods to achieve high sensitivity at low neutron doses. The feasibility was shown of enhancing the PIN diode sensitivity to make possible the measurement of the low doses of neutrons encountered in space flights. The new PIN diode will make possible the development of a very compact, accurate, personal neutron dosimeter

  14. A new electret dosimeter for fast neutrons

    International Nuclear Information System (INIS)

    Campos, L.L.; Suarez, A.A.; Mascarenhas, S.

    1982-01-01

    A new electret for fast-neutron personnel dosimetry is described and calibration curves obtained. Its performance may be improved by changes in the wall composition and geometric parameters. The advantages of electrets over TL and film are the non-erasure of information, low cost, fast reading and portability. (U.K.)

  15. The accreditation programs and proficiency test in Taiwan for personnel dosimeter services (1991-1998)

    International Nuclear Information System (INIS)

    Lee, B.T.; Hwang, W.S.; Su, S.H.

    2000-01-01

    According to the ionizing radiation safety regulation approved by the ROC Atomic Energy Council (AEC), personnel dosimeter processors shall be accredited by the Chinese National Laboratory Accreditation (CNLA) program before offering dosimeter services and shall be accredited every two years (now has been rescheduled to be every three years since 1996). The aim of this program is to secure quality and technical capability of personnel dosimeters, and to provide systematic improvement for the internal calibration and testing laboratory applying for accreditation by means of assessment procedures. The criteria used to evaluate the capability at laboratories are ISO/IEC 25 (1990) and technical guide in different accreditation fields. The Institute of Nuclear Energy Research (INER) was entrusted by CNLA as the central laboratory to perform the proficiency test of personnel dosimeters for laboratory accreditation in December 1990. Such proficiency tests, based on ANSI N13.11 (1983), which is mainly separated to accident categories and protection categories which consists of eight parts of tests including single and mixture radiation of x-ray, gamma, beta and neutron, have been conducted four times, in 1991, 1993, 1995 and 1998. This paper deals with the test procedures and results of proficiency tests of personnel dosimeters from 1991 until 1998. The results of the four proficiency tests showed that, for accident categories, the pass rate is about 91%; for protection categories, the pass rate is about 98%. Meanwhile, the central laboratory will adopt a new version of HPS N13.11 (1993) to replace ANSI N13.11 (1983) as new criteria for the next proficiency test to be conducted in 2001. (author)

  16. Design of calibration method in neutron and individual dosimeter

    International Nuclear Information System (INIS)

    Belkhodia, M.

    1984-12-01

    Usually albedo dosemeters are calibrated with beam of monoenergetic neutrons. Since neutron energy around neutron sources varies greatly, we applied the calibration method to a mixed field whose energy spectrum lies between 0.025 ev and 10 Mev. The method is based on a mathematical model that deals with the dosimeter response as a function at the neutron energy. The measurements carried out with solid state nuclear track detectors show the dosimeter practical aspect. The albedo dosimeter calibration gave results on good agreement with the international institution recommendations

  17. Neutron personnel dosimetry considerations for fusion reactors

    International Nuclear Information System (INIS)

    Barton, T.P.; Easterly, C.E.

    1979-07-01

    The increasing development of fusion reactor technology warrants an evaluation of personnel neutron dosimetry systems to aid in the concurrent development of a radiation protection program. For this reason, current state of knowledge neutron dosimeters have been reviewed with emphasis placed on practical utilization and the problems inherent in each type of dosimetry system. Evaluations of salient parameters such as energy response, latent image instability, and minimum detectable dose equivalent are presented for nuclear emulsion films, track etch techniques, albedo and other thermoluminescent dosimetry techniques, electrical conductivity damage effects, lyoluminescence, thermocurrent, and thermally stimulated exoelectron emission. Brief summaries of dosimetry regulatory requirements and intercomparison study results help to establish compliance and recent trends, respectively. Spectrum modeling data generated by the Neutron Physics Division of Oak Ridge National Laboratory for the Princeton Tokamak Fusion Test Reactor (TFTR) Facility have been analyzed by both International Commission on Radiological Protection fluence to dose conversion factors and an adjoint technique of radiation dosimetry, in an attempt to determine the applicability of current neutron dosimetry systems to deuterium and tritium fusion reactor leakage spectra. Based on the modeling data, a wide range of neutron energies will probably be present in the leakage spectra of the TFTR facility, and no appreciable risk of somatic injury to occupationally exposed workers is expected. The relative dose contributions due to high energy and thermal neutrons indicate that neutron dosimetry will probably not be a serious limitation in the development of fusion power

  18. Personnel neutron dosimetry using TLD elements at PNC

    International Nuclear Information System (INIS)

    Ishiguro, Hideharu

    1985-01-01

    The evaluation method of neutron dose equivalent was studied on the basis of the albedo type neutron dosimetory to design the personnel dosimeter. The dosimeter was composed of three 6 Li 2 10 B 4 O 7 (Cu) TL elements and one 7 Li 2 11 B 4 O 7 (Cu) element. The equations for assessing thermal, epithermal and fast neutron dose equivalents were derived by 252 Cf, 241 Am-Be and PuO 2 neutron sources. The minimum detectable amount of 6 Li 2 10 B 4 O 7 (Cu) element to thermal neutron was 0.02 m rem. The neutron dose equivalent and the gamma one were evaluated separately within about 20 % error in the mixed radiation field. (author)

  19. Description and evaluation of the Hanford personnel dosimeter program from 1944 through 1989. [Contain Glossary

    Energy Technology Data Exchange (ETDEWEB)

    Wilson, R.H.; Fix, J.J.; Baumgartner, W.V.; Nichols, L.L.

    1990-09-01

    This report describes the evolution of personnel dosimeter technology at Hanford since the inception of Hanford operations in 1944. Each of the personnel dosimeter systems used by people working or visiting Hanford is described. In addition, the procedures used to calibrate and calculate dose for each of the dosimeter systems are described. The accuracy of the recorded dose, primarily whole body deep dose, for the different dosimeter systems is evaluated. The evaluation is based on an extensive review of historical literature, as well as a 1989 intercomparison study of all film dosimeters and performance testing of the thermoluminescent dosimeter, also conducted during 1989. 73 refs., 40 figs., 41 tabs.

  20. Neutron dosimetry at commercial nuclear plants. Final report of Subtask B: dosimeter response

    Energy Technology Data Exchange (ETDEWEB)

    Cummings, F.M.; Endres, G.W.R.; Brackenbush, L.W.

    1983-03-01

    As part of a larger program to evaluate personnel neutron dosimetry at commercial nuclear power plants, this study was designed to characterize neutron dosimeter responses inside the containment structure of commercial nuclear plants. In order to characterize those responses, dosimeters were irradiated inside containment at 2 pressurized water reactors and at pipe penetrations outside the biological shield at two boiling water reactors. The reactors were operating at full power during the irradiations. Measurements were also performed with electronic instruments, the tissue equivalent proportional counter (TEPC), and portable remmeters, SNOOPY, RASCAL and PNR-4.

  1. Neutron dosimetry at commercial nuclear plants. Final report of Subtask B: dosimeter response

    International Nuclear Information System (INIS)

    Cummings, F.M.; Endres, G.W.R.; Brackenbush, L.W.

    1983-03-01

    As part of a larger program to evaluate personnel neutron dosimetry at commercial nuclear power plants, this study was designed to characterize neutron dosimeter responses inside the containment structure of commercial nuclear plants. In order to characterize those responses, dosimeters were irradiated inside containment at 2 pressurized water reactors and at pipe penetrations outside the biological shield at two boiling water reactors. The reactors were operating at full power during the irradiations. Measurements were also performed with electronic instruments, the tissue equivalent proportional counter (TEPC), and portable remmeters, SNOOPY, RASCAL and PNR-4

  2. Angular dependence of dose equivalent response of an albedo neutron dosimeter

    International Nuclear Information System (INIS)

    Torres, B.A.; Boswell, E.; Schwartz, R.B.

    1994-01-01

    The ANSI provides procedures for testing the performance of dosimetry services. Although neutron dose equivalent angular response studies are not now mandated, future standards may well require that such studies be performed. Current studies with an albedo dosimeter will yield information regarding the angular dependence of dose equivalent response for this type of personnel dosimeter. Preliminary data for bare 252 Cf fluences show a marked decrease in dosimeter reading with increasing angle. The response decreased by an approximate factor of four. For the horizontal orientation, the same response was noted from both positive and negative angles. However, for the vertical orientation, the response was unexplainably assymetric. We are also examining the response of the personnel badge in moderated 252 Cf fluences. Responses from the moderated and unmoderated 252 Cf fields and theoretical calculations of the neutron angular response will be compared. This information will assist in building a data base for future comparisons of neutron angular responses with other neutron albedo dosimeters and phantoms

  3. Design, construction and characterization of a dosimeter for neutron radiation

    International Nuclear Information System (INIS)

    Souto, Eduardo de Brito

    2007-01-01

    An individual dosimeter for neutron-gamma mixed field dosimetry was design and developed aiming monitoring the increasing number of workers potentially exposed to neutrons. The proposed dosimeter was characterized to an Americium-Beryllium source spectrum and dose range of radiation protection interest (up to 20 mSv). Thermoluminescent albedo dosimetry and nuclear tracks dosimetry, traditional techniques found in the international literature, with materials of low cost and national production, were used. A commercial polycarbonate, named SS-1, was characterized for solid state tack detector application. The chemical etching parameters and the methodology of detectors evaluation were determined. The response of TLD-600, TLD-700 and SS-1 were studied and algorithms for dose calculation of neutron and gamma radiation of Americium- Beryllium sources were proposed. The ratio between thermal, albedo and fast neutrons responses, allows analyzing the spectrum to which the dosimeter was submitted and correcting the track detector response to variations in the radiation incidence angle. The new dosimeter is fully characterized, having sufficient performance to be applied as neutron dosimeter in Brazil. (author)

  4. Intercomparison of personnel dosimetry for thermal neutron dose equivalent in neutron and gamma-ray mixed fields

    International Nuclear Information System (INIS)

    Ogawa, Yoshihiro

    1985-01-01

    In order to consider the problems concerned with personnel dosimetry using film badges and TLDs, an intercomparison of personnel dosimetry, especially dose equivalent responses of personnel dosimeters to thermal neutron, was carried out in five different neutron and gamma-ray mixed fields at KUR and UTR-KINKI from the practical point of view. For the estimation of thermal neutron dose equivalent, it may be concluded that each personnel dosimeter has good performances in the precision, that is, the standard deviations in the measured values by individual dosimeter were within 24 %, and the dose equivalent responses to thermal neutron were almost independent on cadmium ratio and gamma-ray contamination. However, the relative thermal neutron dose equivalent of individual dosimeter normalized to the ICRP recommended value varied considerably and a difference of about 4 times was observed among the dosimeters. From the results obtained, it is suggested that the standardization of calibration factors and procedures is required from the practical point of radiation protection and safety. (author)

  5. Evaluation of spurious readings in Los Alamos personnel TL dosimeters

    International Nuclear Information System (INIS)

    Eisen, Y.; Littlejohn, G.J.; Cortez, J.R.

    1984-08-01

    This study investigates the possibility of tritium build-up in TLD-600 chips irradiated with neutrons and the causes of spurious readings in the Harshaw TLD cards used for personnel dosimetry. Experiments indicated that spurious readings in TLD-600 chips, previously irradiated with neutrons, might occur in cases where the cards had been accidentally read at temperatures lower than 300 0 C as a result of bad contact between the hot finger in the reading system and the chips. Because a TLD card contains glue and paper bar-code labels, the postannealing is performed at 80 0 C for 17 hours. This annealing procedure alone does not effectively deplete the high-energy traps, such as those near 260 0 C, populated by high-LET (Linear Energy Transfer) particles. TLD-600 chips, irradiated on a phantom by 400 mrem of moderated fission neutrons, read at 240 0 C, annealed at 80 0 C for 17 hours, and then reread at 280 0 C, showed residual doses as large as 200 mrem (equivalent photons). Calculations and experiments show that for neutron exposures around 1 rem of moderated fission neutrons with an average energy of 500 keV, the maximum build-up of dose as a result of tritium formation is less than 1 mrem. The dose build-up in properly annealed TLD-600 and TLD-700 chips, is nearly the same, even though the TLD-600 chips were previously irradiated by neutrons. Both kinds of chips show natural background accumulation. A mechanism for annealing the Harshaw cards at high temperatures, without destroying the label or the adhesive material, was developed and found to be useful. 7 references, 4 figures, 3 tables

  6. Investigations of CR39 dosimeters for neutron routine dosimetry

    International Nuclear Information System (INIS)

    Weinstein, M.; Abraham, A.; Tshuva, A.; German, U.

    2004-01-01

    CR-39 is a polymeric nuclear track detector which is widely used for neutron dosimetry. CR-39 detector development was conducted at a number of laboratories throughout the world(1,2) , and was accepted also for routine dosimetry. However, there are shortcomings which must be taken into consideration the lack of a dosimetry grade material which causes batch variations, significant angular dependence and a moderate sensitivity. CR-39 also under-responds for certain classes of neutron spectra (lower energy neutrons from reactors or high energy accelerator-produced neutrons).In order to introduce CR-39 as a routine dosimeter at NRCN, a series of checks were performed. The present work describes the results of some of our checks, to characterize the main properties of CR-39 dosimeters

  7. The Martin Marietta Energy Systems personnel neutron dosimetry program

    International Nuclear Information System (INIS)

    McMahan, K.L.

    1991-01-01

    Martin Marietta Energy Systems, Inc. (Energy Systems), manages five sites for the US Department of Energy. Personnel dosimetry for four of the five sites is coordinated through a Centralized External Dosimetry System (CEDS). These four sites are the Oak Ridge National Laboratory (ORNL), the Oak Ridge Y-12 Plant (Y-12), the Oak Ridge K-25 Site (K-25), and the Paducah Gaseous Diffusion Plant (PGDP). The fifth Energy Systems site, Portsmouth Gaseous Diffusion Plant, has an independent personnel dosimetry program. The current CEDS personnel neutron dosimeter was first issued in January 1989, after an evaluation and characterization of the dosimeters' response in the workplaces was performed. For the workplace characterization, Energy Systems contracted with Pacific Northwest Laboratory (PNL) to perform neutron measurements at selected locations at ORNL and Y-12. K-25 and PGDP were not included because their neutron radiation fields were similar to others already planned for characterization at ORNL and Y-12. Since the initial characterization, PNL has returned to Oak Ridge twice to perform follow up measurements, and another visit is planned in the near future

  8. SEVENTH DOE WORKSHOP ON PERSONNEL NEUTRON DOSIMETRY

    Energy Technology Data Exchange (ETDEWEB)

    Vallario, E J

    1978-10-24

    This workshop was the seventh of a series and was held on October 23-24. 1978, at the Central Electricity Generating Board, HQ, London, England. Typically~ attendees at the Workshop were concerned with one of three activities: studying and refining existing techniques in an attempt to quantify already-known parameters with greater precision, looking for ways to apply existing neutron dosirr:etry techniques to a specific local problem, identifying the needs and weaknesses of existing systems, with the goal of improving and passibly simplifying field measurements. The types of neutron dosimetry techniques discussed by participants included albedo dosimeters, track etch, and TLD. One speaker reported on NTA film, noting that fading could be eliminated by drying the emulsion in dry nitrogen before field use. There were no reports on tissue equivalent proportional counters or activation analysis. One participant discussed a metal oxide silicon dosimeter. The need to develop a consistent standard terminology, as well as calibration sources and techniques, on both the national and international level was evident. The need for standardization is particularly acute in the U.S. Techniques for evaluating dosimeter response in the field should he standardized, since several different instruments with widely different response characteristics are currently being used. The choice of instruments is often parochial. Also. the type and use of phantoms should be standardized. Neutron dose assignment is significantly affected by the position of the dosimeter on the body. for example, a typical albedo dosimeter may give differences of up to 20% depending on whether it is worn on the belt or chest. Larger errors are encountered with front-to-back (angular} orientation. 1n an attempt to minimize such errors~ at least two European facilities are using neutron dosimeter belts, which provide dosimeters both in front and in back of the wearer. The gamma-to-neutron ratio around nuclear power

  9. Automatic scanning of NTA film neutron dosimeters

    CERN Document Server

    Müller, R

    1999-01-01

    At the European Laboratory for Particle Physics CERN, personal neutron monitoring for over 4000 collaborators is performed with Kodak NTA film, one of the few suitable dosemeters in the stray radiation environment of a high energy accelerator. After development, films are scanned with a projection microscope. To overcome this lengthy and strenuous procedure an automated analysis system for the dosemeters has been developed. General purpose image recognition software, tailored to the specific needs with a macro language, analyses the digitised microscope image. This paper reports on the successful automatic scanning of NTA films irradiated with neutrons from a /sup 238/Pu-Be source (E approximately=4 MeV), as well as on the extension of the method to neutrons of higher energies. The question of detection limits is discussed in the light of an application of the method in routine personal neutron monitoring. (9 refs).

  10. Performance of neutron and gamma personnel dosimetry in mixed radiation fields

    International Nuclear Information System (INIS)

    Swaja, R.E.; Sims, C.S.

    1981-01-01

    From 1974 to 1980, six personnel dosimetry intercomparison studies (PDIS) were conducted at the Oak Ridge National Laboratory (ORNL) to evaluate the performance of personnel dosimeters in a variety of neutron and gamma fields produced by operating the Health Physics Research Reactor (HPRR) in the steady state mode with and without spectral modifying shields. A total of 58 different organizations participated in these studies which produced approximately 2000 measurements of neutron and gamma dose equivalents on anthropomorphic phantoms for five different reactor spectra. Based on these data, the relative performance of three basic types of neutron dosimeters [nuclear emulsion film, thermoluminescent (TLD), and track-etch] and two basic types of gamma dosimeters (film and TLD) in mixed radiation fields was assessed

  11. The Response of Alanine Dosimeters in Thermal Neutron Fields

    DEFF Research Database (Denmark)

    Schmitz, T.; Bassler, Niels; Sharpe, P.

    response of all pellets could be reproduced by calculations within a uncertainty of 5 %. For all experiments three dose components have been separated. A proton dose is generated in the 14N(n,p)14C reaction. Secondary gammas are generated by various (n,γ) reactions, dominated by the 2.2 MeV photon from...... experiments the dosimeters will be exposed to higher neutron energies, which are more typical for BNCT treatments. References: [1] Barth, R.F; 2009: Boron neutron capture therapy at the crossroads: Challenges and opportunities. Applied Radiation and Isotopes 67, 3-6. [2] Rogus, R.D.; Harling, O.K.; Yanch, J.C...... for treatment of liver metastases. Applied Radiation and Isotopes 67, 238-241. [4] Sharpe, P.; Sephtan, J.; 2000: An automated system for the measurement of alanine/EPR dosimeters. Applied Radiation and Isotopes 52, 1185-1188....

  12. Neutron accident dosimeter based on SSNTDs

    International Nuclear Information System (INIS)

    Palfalvi, J.; Sajo-Bohus, L.

    1998-01-01

    A sandwich type track etch detector of CR-39 was developed utilizing neutron-proton recoil and (n,α) reactions. Applying gold and Cd filters this system turns into a threshold detector and also it combines the albedo and the direct detection methods; thus it becomes possible to detect neutrons in three or more energy ranges depending on the number of gold degraders of different thickness allowing dose assessment with an uncertainty of about 20%, as blank tests have proved when a single gold foil of 20 μm thick was used. (author)

  13. Personnel radiation dosimetry laboratory accreditation programme for thermoluminescent dosimeters : a proposal

    International Nuclear Information System (INIS)

    Bhatt, B.C.; Srivastava, J.K.; Iyer, P.S.; Venkatraman, G.

    1993-01-01

    Accreditation for thermoluminescent dosimeters is the process of evaluating a programme intending to use TL personnel dosimeters to measure, report and record dose equivalents received by radiation workers. In order to test the technical competence for conducting personnel dosimetry service as well as to decentralize personnel monitoring service, it has been proposed by Radiological Physics Division (RPhD) to accredit some of the laboratories, in the country. The objectives of this accreditation programme are: (i) to give recognition to competent dosimetry processors, and (ii) to provide periodic evaluation of dosimetry processors, including review of internal quality assurance programme to improve the quality of personnel dosimetry processing. The scientific support for the accreditation programme will be provided by the scientific staff from Radiological Physics Division (RPhD) and Radiation Protection Services Division (RPSD). This paper describes operational and technical requirements for the Personnel Radiation Dosimetry Laboratory Accreditation Programme for Thermoluminescent Dosimeters for Personnel Dosimetry Processors. Besides, many technical documents dealing with the TL Personnel Dosimeter System have been prepared. (author). 5 refs., 2 figs

  14. An automatic evaluation system for NTA film neutron dosimeters

    CERN Document Server

    Müller, R

    1999-01-01

    At CERN, neutron personal monitoring for over 4000 collaborators is performed with Kodak NTA films, which have been shown to be the most suitable neutron dosimeter in the radiation environment around high-energy accelerators. To overcome the lengthy and strenuous manual scanning process with an optical microscope, an automatic analysis system has been developed. We report on the successful automatic scanning of NTA films irradiated with sup 2 sup 3 sup 8 Pu-Be source neutrons, which results in densely ionised recoil tracks, as well as on the extension of the method to higher energy neutrons causing sparse and fragmentary tracks. The application of the method in routine personal monitoring is discussed. $9 overcome the lengthy and strenuous manual scanning process with an optical microscope, an automatic analysis system has been developed. We report on the successful automatic scanning of NTA films irradiated with /sup 238/Pu-Be source $9 discussed. (10 refs).

  15. Development and characterization of real-time wide-energy range personal neutron dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Nakamura, Takashi; Tsujimura, Norio (Tohoku Univ., Sendai (Japan). Cyclotron and Radioisotope Center); Yamano, Toshiya; Suzuki, Toshikazu; Okamoto, Eisuke

    1994-04-01

    The authors developed a real-time personal neutron dosimeter which could give neutron dose equivalent over wide energy region from thermal to 10 odd MeV by using 2 silicon detectors, fast neutron sensor and slow neutron sensor. The energy response of this dosimeter was evaluated under thermal neutron field, monoenergetic neutron field between 200 keV and 15 MeV, and moderated [sup 252]Cf neutron field. The neutron dose equivalent was estimated by adding neutron dose equivalent below 1 MeV given by slow neutron sensor and that above 1 MeV by fast neutron sensor. It was verified from various field tests that this dosimeter is able to give neutron dose equivalent within a factor of 2 margin of accuracy in reactor, accelerator, fusion research and nuclear fuel handling facilities. This dosimeter has more than one order higher sensitivity than conventional personal neutron dosimeters and is insensitive to [gamma]-rays up to about 500 mSv/h. This dosimeter will soon be commercially available as a personal dosimeter which gives neutron and [gamma]-ray dose equivalents simultaneously by installing [gamma]-ray silicon sensor. (author).

  16. Personnel monitoring system by thermoluminescence dosimeter - a study and review

    International Nuclear Information System (INIS)

    Iyer, V.S.; Alok, M.; Chandel, B.S.

    2018-01-01

    The principle objective of personnel monitoring system is to prevent overexposure and avoid unnecessary exposure of personnel working with various sources of radiation. The adequate system of personnel monitoring must provide for the measurement. evaluation and recording of significant doses accumulated by individuals together with a recording of conditions under which these doses are received

  17. Determination of personnel exposures in the lower energy ranges of X-ray by photographic dosimeter

    International Nuclear Information System (INIS)

    Ha, C.W.; Kim, J.R.; Suk, K.W.

    1986-01-01

    This paper described an improved technical method required for proper evaluation of personnel exposures by means of the photographic dosimeter developed by KAERI in lower gamma or X-ray energy regions, with which response of the dosimeter varies significantly. With calibration of the dosimeter in the energy range from 30 to 300 keV, the beam spectrum was carefully selected and specified it adequately. The absorber combinations and absorber thickness used to obtain the specified X-ray spectra from a constant potential X-ray machine were determined theoretically and also experimentally. A correlation between the density and exposure for the four separate energies, such as 49 keV eff , 154 keV eff 250 keV eff and 662 keV, is experimentally determined. As a result, it can be directly evaluated the exposure from the measured response of dosimeter. (Author)

  18. Detector and dosimeter for neutrons and other radiation

    International Nuclear Information System (INIS)

    Apfel, R.E.

    1979-01-01

    A radiation detector and dosimeter is based on the fact that a sufficiently finely-dispersed liquid suspended in a host liquid of high viscosity or gel is stable at temperatures above its normal boiling point for long periods of time provided it is protected from contact with walls, or other types of initiators which can cause volatilization or vaporization of the droplets. Radiation, and particularly neutron radiation of sufficient energy and intensity on coming in contact with such droplets can trigger volatilization. The volume of vapor evolved can then serve as a measure of radiation intensity and dosage

  19. Development of neutron dosimeter using CR-39 for measurement of ambient dose equivalent

    International Nuclear Information System (INIS)

    Maki, Daisuke; Shinozaki, Wakako; Ohguchi, Hiroyuki; Yamamoto, Takayoshi; Nakamura, Takayoshi

    2010-01-01

    A CR-39 has good advantages such as cumulative type dosimeter, small fading effect and gamma-ray insensitive. Therefore, we developed the wide energy-range environmental neutron dosimeter using eight CR-39s for area monitoring in this study. This dosimeter is made of octagonal columnar polyethylene block which height is 60 mm and bottom side is 25 mm. The dosimeter contains two types of CR-39s for fast neutron detection and slow neutron detection. Four CR-39s for fast neutron detection are used for detection of recoil protons produced by H (n, p) reactions. Four CR-39s for slow neutron detection are used with boron nitride converter to detect alpha-rays produced by 10 B (n, α) 7 Li reactions. Ambient dose equivalent is obtained by adding the number of etch-pits observed in four CR-39s for fast neutron detection to the number of etch-pits observed in four CR-39s for slow neutron detection with appropriate constants respectively. Dosimeters were irradiated with some energetic neutrons and evaluated results of ambient dose equivalent were compared with results from neutron transport calculations. Energy response of dosimeter shows good agreement with neutron fluence to ambient dose equivalent conversion coefficients. Directional dependence of dosimeter is at the same level as the rem-counter. (author)

  20. Triton and alpha-particle contribution from LiF converter for neutron dosimeter

    CERN Document Server

    Camacho, M E; Balcazar, M

    1999-01-01

    A personnel neutron dosimeter prototype based on chemical and electrochemical etched CR-39 detector, combined with LiF converter, has been calibrated using an ICRP-like phantom, under a heavy-water moderated Californium source neutron spectra; A conversion factor of 1.052+-126 spots cm sup - sup 2 mSv sup - sup 1 was obtained. The sealing properties of the detector holder showed a ten-fold reduction in radon background when it was tested in a high radon atmosphere. A convenient mechanical shock resistance was achieved in LiF converters by sintering to 11 tons pressure LiF powder at 650 deg. C, during one hour.

  1. Personnel neutron dose assessment upgrade: Volume 2, Field neutron spectrometer for health physics applications

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Reece, W.D.; Miller, S.D.

    1988-07-01

    Both the (ICRP) and the (NCPR) have recommended an increase in neutron quality factors and the adoption of effective dose equivalent methods. The series of reports entitled Personnel Neutron Dose Assessment Upgrade (PNL-6620) addresses these changes. Volume 1 in this series of reports (Personnel Neutron Dosimetry Assessment) provided guidance on the characteristics, use, and calibration of personnel neutron dosimeters in order to meet the new recommendations. This report, Volume 2: Field Neutron Spectrometer for Health Physics Applications describes the development of a portable field spectrometer which can be set up for use in a few minutes by a single person. The field spectrometer described herein represents a significant advance in improving the accuracy of neutron dose assessment. It permits an immediate analysis of the energy spectral distribution associated with the radiation from which neutron quality factor can be determined. It is now possible to depart from the use of maximum Q by determining and realistically applying a lower Q based on spectral data. The field spectrometer is made up of two modules: a detector module with built-in electronics and an analysis module with a IBM PC/reg sign/-compatible computer to control the data acquisition and analysis of data in the field. The unit is simple enough to allow the operator to perform spectral measurements with minimal training. The instrument is intended for use in steady-state radiation fields with neutrons energies covering the fission spectrum range. The prototype field spectrometer has been field tested in plutonium processing facilities, and has been proven to operate satisfactorily. The prototype field spectrometer uses a 3 He proportional counter to measure the neutron energy spectrum between 50 keV and 5 MeV and a tissue equivalent proportional counter (TEPC) to measure absorbed neutron dose

  2. Fast neutron personnel dosimetry by CR-39 plastics a new electrochemical etching procedure

    International Nuclear Information System (INIS)

    Djeffal, S.

    1984-07-01

    In the first part of this work a brief description of solid state nuclear track detectors, the principles of track registration and the different reading techniques are given. In the experimental part of the present work we systematically analysed different etching procedures and set a new electrochemical etching method, which enables us to develop a new fast neutron dosimeter. This fast neutron dosimeter makes possible the measurement of low neutron doses in the energy range from 10 Kev to 20 Mev with a reasonably flat energy response. These new developments are very attractive in personnel neutron dosimetry where nuclear emulsions are still used despite their insensitivity to neutron energies down to 500 Kev (i.e. the energy range one often encounters around nuclear facilities)

  3. Angular response characterization of the Martin Marietta Energy Systems, Inc., personnel dosimeter

    International Nuclear Information System (INIS)

    Ahmed, A.B.; McMahan, K.L.; Colwell, D.S.

    1993-08-01

    An evaluation of the Martin Marietta Energy Systems, Inc., personnel dosimeter to radiation incident from non-perpendicular angles was carried out to meet the Department of Energy Laboratory Accreditation Program (DOELAP) requirements. Dosimeters were exposed to six different radiation sources. For each source, dosimeters were rotated about their horizontal and vertical axes at seven different angles each. Raw readings were processed through the dose calculation algorithm used for routine personnel dosimetry to determine dose equivalent values. Dose equivalent responses relative to zero degree incident angle were found to be within ± 20% for M150, K-59 and 137 Cs photons when the incident angle was 60 degree or less. For low-energy photon irradiations (M30 and K-16), responses for angles other than perpendicular incidence are generally unpredictable. Reasons include: (1) failure of dose calculation algorithm to identify the radiation field correctly due to unusual element ratios; and (2) at extreme angles (± 85 degree), the dosimeter design (in relation to the irradiation geometry) becomes the limiting factor in producing reproducible results. Response to 204 Tl beta particles decreases rapidly with increasing angle of incidence

  4. Evaluation of the implementation and use of active personal dosimeters for neutrons in Brazil

    International Nuclear Information System (INIS)

    Castro B, C. P.; Wagner P, W.; De Souza P, K. C.

    2014-08-01

    This work was conducted through of a field research based on a questionnaire sent to users of active personal dosimeters. A retrospective study of the last six years was also carried out of the services in the Neutron Metrology Laboratory (2008-2013) referent to the active personal dosimeters, taking into consideration the standards ISO-8529-3 and IEC-61526. The active personal dosimeters are defined as any instrument of individual monitoring with direct reading capacity, used by individuals exposed to ionizing radiation fields. Through research was verified that the active personal dosimeters work associated with other dosimeter types. Considering all dosimeters declared in the questionnaire, only two dosimeters (MGP brand Dmc 2000-GN model and the brand ATOMTEX model AT2503A) have conformity declaration with the international standard IEC-61526: 2005 reported by the manufacturers. (author)

  5. Evaluation of a thermoluminescent dosimeter for personnel monitoring in the nuclear-radiation environment

    International Nuclear Information System (INIS)

    Johnson, F.A.; McGowan, S.; Gravelle, R.A.

    1983-09-01

    Efforts to improve the responses of phosphor/polyethylene TLDs to neutrons in the 1-MeV range have been only partially successful. Some increases in response have resulted from reduction of phosphor grain size and substitution of CaSO 4 :Tm with CaF 2 :Mn or CaF 2 :Dy. Reasons for the continuing low response are not fully understood. Calculated responses of the TLDs, when located on the body, show a closer correlatin with the dose received by the bone marrow as a result of exposure to a weapon spectrum than is evident for the case of flat-response dosimeters. However the TLDs exhibit a greater difference in their response to a weapon spectrum as compared with that to a pure gamma spectrum than is the case for a flat-response dosimeter. The disparity becomes even larger when the dose measured by the dosimeters is expressed in terms of the biologically effective dose absorbed in the bone marrow. Investigation of the ability of a dosimeter, which is sensitive only to γ-rays, to provide a measure of the total bone marrow dose produced by a combined flux of γ-rays and neutrons, indicates that there is an unacceptable difference between the response to a pure gamma spectrum and that to a mixed spectrum from a weapon

  6. Neutron dosimeters and survey meters in accelerators, reactors and other neutron environments

    International Nuclear Information System (INIS)

    1989-03-01

    Neutron fields in occupationally accessible areas around nuclear reactors, radioisotope sources and medical and high energy accelerators have been characterized using currently available information. Neutron, rem meters, such as the Leake detector, are the most suitable instruments available for conducting neutron dose rate surveys in the vicinity of radioisotope neutron sources, nuclear reactors and medical accelerators. However, these instruments have been shown to be deficient in that they overrespond by a factor of four to neutrons in the 0.1 to 1 MeV range and are insensitive to neutrons from about 1 eV up to about 10 keV. Also, they are insensitive to neutrons above 20 MeV and their use must be restricted near high energy accelerators where significant numbers of neutrons above 20 MeV are known to be present. The most suitable instrument of measure dose from neutrons above 20 MeV is the 12 C(n,2n) 11 C scintillation chamber. Commercially available rem meters frequently use BF 3 counters in the pulse mode to detect thermal neutrons. Therefore, measurements around pulsed accelerators must be made with caution to ensure that the detector is not saturated during each pulse and that the accelerator pulse period is greater than the response time of the detector. The personal neutron dosimeters currently available either are known to be insensitive to neutrons above 20 MeV or have not been tested. Also, all except the albedo dosimeter are insensitive to or have not been tested for neutron energies in the range 1 eV to 10 keV. Several dosimeter types respond reasonably well to neutrons in the energy range 0.1 to 15 MeV, for example, CR-39, bubble and superheated drop detectors. However, the first gas a lower limit of sensitivity of about 0.3 mSv. The bubble detector can be designed to measure doses as small as 1μSv and offers the additional benefit of direct-reading capability. The superheated drop detector is not suitable for use around pulsed accelerators because

  7. A quality control program for the thermoluminescence dosimeter (TLD) in personnel radiation monitoring

    International Nuclear Information System (INIS)

    Seo, Kyung Won; Kim, Jang Lyul; Lee, Sang Yoon; Lee, Hyung Sub

    1994-01-01

    High quality radiation dosimetry is essential for workers who rely upon personal dosemeters to record the amount of radiation to which they are exposed. The ministry of science and technology (MOST) issued a ministerial ordinance (No 199-15) about the technical criteria on personnel dosimeter processors on 1992. The purpose of this quality control program is to prescribe the procedures approved by the management of KAERI for implementing a quality badge service by means of TLD for personnel working in an area where they may be exposed to ionization radiation. (Author) 10 refs

  8. Personnel dose equivalent monitoring at SLAC using lithium-fluoride TLD's [thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Jenkins, T.M.; Busick, D.D.

    1987-03-01

    TLD's replaced film badges in the early 1970's for all dose equivalent monitoring, both neutron and photon, and for all locations at SLAC. The photon TLD's, composed of Li-7 loaded teflon discs, are calibrated using conventional gamma-ray sources; i.e., Co-60, Cs-137, etc. For these TLD's a nominal value of 1 nC/mrem is used, and is independent of source energy for 100 keV to 3 MeV. Since measured dose equivalents at SLAC are only a small fraction of the allowable levels, it was not deemed necessary to develop neutron dosimeters which would measure dose equivalent accurately for all possible neutron spectra. Today, wallet TLD's, composed of pairs of Li-7 and Li-6 discs, are used, with the Li-6 measuring only thermal neutrons; i.e., they aren't moderated in any way to make them sensitive to neutrons with energies greater than thermal. The assumption is made that there is a correlation between thermal neutron fluences and fast neutron fluences around the research area where almost all neutron doses (exclusive of sealed sources) are received. The calibration factor for these Li-6 TLD's is 1 nC/mrem of fast neutrons. The method of determining the validity of this calibration is the subject of this note. 4 refs., 9 figs., 1 tab

  9. Dosimeter

    International Nuclear Information System (INIS)

    Thomson, I.

    1986-01-01

    This invention relates to a dosimeter for measuring ionizing radiation, and particularly to a dosimeter using an insulated gate field effect transistor (IGFET) as a sensor, having substantially improved accuracy. An IGFET is a field effect transistor fabricated on a silicon substrate and having an oxide insulator between the gate electrode and the silicon substrate. The gate electrode can be either metal or polycrystalline silicon dioxide. This invention overcomes previously-noted problems with IGFET sensors - the variation of threshold voltage with temperature, their inherent zero offset which varies from wafer to wafer, and the zero drift in threshold voltage - by measuring the differential threshold between two IGFET sensors exposed to the same radiation, in which one is biased into its conducting region, and the other is biased either off or to a conducting level less than the first. The measured differential threshold voltage between the two transistors will be a measure of the gamma radiation dose

  10. Development of real time personal neutron dosimeter with two silicon detectors

    Energy Technology Data Exchange (ETDEWEB)

    Nakamura, T.; Tsujimura, N. [Tohoku Univ., Cyclotron and Radioisotope Center, Aoba, Aramaki, Aoba-ku (Japan); Yamano, T. [Tokyo Factory, Fuji Electric Co. Ltd., Tokyo (Japan)

    1992-07-01

    We developed a real time personal neutron dosimeter by using two types of silicon p-n junction detectors, thermal neutron sensor and fast neutron sensor. The thermal neutron sensor which is {sup 10}B doped n-type silicon with a polyethylene radiator mainly counts neutrons of energy front thermal to I MeV, and the fast neutron sensor which is p-type silicon with a polyethylene radiator is sensitive to neutrons above I MeV. The neutron sensitivity measurements revealed that the dosimeter has a rather flat response for dose equivalent from thermal to 15 MeV, excluding a drop from 50 keV to I MeV. In order to get conversion factor from counts to dose equivalent as accurately as possible, we performed the field test of the dosimeter calibration in several neutron-generating fields. By introducing the two-group dose estimation method, this dosimeter can give the neutron dose equivalent within about 50% errors. (author)

  11. Automated gamma spectrometry and data analysis on radiometric neutron dosimeters

    International Nuclear Information System (INIS)

    Matsumoto, W.Y.

    1983-01-01

    An automated gamma-ray spectrometry system was designed and implemented by the Westinghouse Hanford Company at the Hanford Engineering Development Laboratory (HEDL) to analyze radiometric neutron dosimeters. Unattended, automatic, 24 hour/day, 7 day/week operation with online data analysis and mainframe-computer compatible magnetic tape output are system features. The system was used to analyze most of the 4000-plus radiometric monitors (RM's) from extensive reactor characterization tests during startup and initial operation of th Fast Flux Test Facility (FFTF). The FFTF, operated by HEDL for the Department of Energy, incorporates a 400 MW(th) sodium-cooled fast reactor. Aumomated system hardware consists of a high purity germanium detector, a computerized multichannel analyzer data acquisition system (Nuclear Data, Inc. Model 6620) with two dual 2.5 Mbyte magnetic disk drives plus two 10.5 inch reel magnetic tape units for mass storage of programs/data and an automated Sample Changer-Positioner (ASC-P) run with a programmable controller. The ASC-P has a 200 sample capacity and 12 calibrated counting (analysis) positions ranging from 6 inches (15 cm) to more than 20 feet (6.1 m) from the detector. The system software was programmed in Fortran at HEDL, except for the Nuclear Data, Inc. Peak Search and Analysis Program and Disk Operating System (MIDAS+)

  12. Sensitive silicon PIN-diode dosimeter for fast neutrons and method to control and increase its sensitivity

    International Nuclear Information System (INIS)

    Swinehart, P.R.; Swartz, J.M.

    1978-01-01

    With the personnel dosimeter, applicable e.g. in medicine, a dose of 0.1 rad for neutrons with an energy greater than 10 keV can be detected. In the range between 0.1 and 20 rad sensitivity is increased to 5 mV/rad. This sensitivity can be achieved by distributing the mass of the semiconductor material of the diode or equal to four times the reciprocal base width. Appropriate dimensions are 750 μm for the edge length of the end surface resp. diameter and 750 μm up to 5000 μm for the base width. (DG) [de

  13. DOE personnel neutron dosimetry evaluation and upgrade program

    International Nuclear Information System (INIS)

    Faust, L.G.; Stroud, C.M.; Vallario, E.J.

    1988-01-01

    The US Department of Energy (DOE) sponsors an extensive research program to improve the methods, dosimeters, and instruments available to DOE facilities for measuring neutron dose and assessing its effects on the work force. The Total Dose Meter was recently developed for measuring in real time the absorbed dose of mixed neutron and gamma radiation and for calculating the dose equivalent. The Field Neutron Spectrometer was developed to provide a portable instrument for determining neutron spectra in the workplace for flux-to-dose equivalent conversion and quality factor calculation. The Combination Thermoluminescence/Track Etch Dosimeter (TLD/TED) was developed to extend the effective neutron energy range of the conventional TLDs to improve detection of fast-energy neutrons. An Optically Stimulated Luminescence Dosimeter is presently being developed for application to gamma, neutron, and beta radiation. An Effective Dose Equivalent System is being developed to provide guidance in implementing the January 1987 Presidential Directive to determine effective dose equivalent. Superheated Drop Detectors are being investigated for their potential as real time neutron dosimeters. This paper includes discussions of these improvements brought about by the DOE research program

  14. Experimental Validation of Ex-Vessel Neutron Spectrum by Means of Dosimeter Materials Activation Method

    Directory of Open Access Journals (Sweden)

    S.A. Santa

    2017-06-01

    Full Text Available Neutron spectrum information in reactor core and around of ex-vessel reactor needs to be known with a certain degree of accuracy to support the development of fuels, materials, and other components. The most common method to determine neutron spectra is by utilizing the radioactivation of dosimeter materials. This report presents the evaluation of neutron flux incident on M3dosimeter sets which were irradiated outside the reactor vessel,as well as the validation of  neutron spectrum calculation. Al capsules containing both dosimeter set covered withCd and dosimeter set without Cd cover have been irradiated during the 35th operational cycle in the M3 ex-vessel irradiation hole position207 cmfrom core centerline at the space between the reactor vessel and the safety vessel. The capsules were positioned at Z=0.0 cm of core midplane. Each dosimeter set consists of Co-Al, Sc, Fe, Np, Nb, Ni, B, and Ta. The gamma-ray spectra of irradiated dosimeter materials were measured by 63 cc HPGe solid-state detector and photo-peak spectra were analyzed using BOB75 code. The reaction rates of each dosimeter materials and its uncertainty were analyzed based on 59Co (n,g 60Co, 237Np (n,f 95Zr-103Ru,  45Sc (n,g 46Sc, 58Fe (n,g 59Fe, 181Ta (n,g 182Ta, and 58Ni (n,p58Co reactions. The measured Cd ratios indicate that neutron spectrum at the irradiated dosimeter sets was dominated by low energy neutron. The experimental result shows that the calculated neutron spectra by DORT code at the ex-vessel positions need correction, especially in the fast neutron energy region, so as to obtain reasonable unfolding result consistent with the reaction rate measurement without any exception. Using biased DORT initial spectrum, the neutron spectrum and its integral quantity were unfolded by NEUPAC code. The result shows that total neutron flux, flux above 1.0 MeV, flux above 0.1 MeV, and the displacement rate of the dosimeter set not covered with Cd were 1.75× 1012 n cm2 s-1, 1

  15. Passive dosimeters other than film and TLDs [thermoluminescent dosimeter

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1986-01-01

    This presentation will describe CR-39 plastic as a personnel neutron dosimeter. Recent research at LLNL and elsewhere has resulted in the development of a dosimetry system that is superior to any personnel neutron dosimeter previously available. The author describes the features of the dosimetry system and the new etching procedures and techniques in detail. Most of the research was done at the LLNL and has been supported as a part of the DOE Neutron Dosimetry Upgrade Program. 10 refs., 4 figs., 1 tab

  16. Investigation of the response characteristics of OSL albedo neutron dosimeters in a 241AmBe reference neutron field

    Science.gov (United States)

    Liamsuwan, T.; Wonglee, S.; Channuie, J.; Esoa, J.; Monthonwattana, S.

    2017-06-01

    The objective of this work was to systematically investigate the response characteristics of optically stimulated luminescence Albedo neutron (OSLN) dosimeters to ensure reliable personal dosimetry service provided by Thailand Institute of Nuclear Technology (TINT). Several batches of InLight® OSLN dosimeters were irradiated in a reference neutron field generated by the in-house 241AmBe neutron irradiator. The OSL signals were typically measured 24 hours after irradiation using the InLight® Auto 200 Reader. Based on known values of delivered neutron dose equivalent, the reading correction factor to be used by the reader was evaluated. Subsequently, batch homogeneity, dose linearity, lower limit of detection and fading of the OSLN dosimeters were examined. Batch homogeneity was evaluated to be 0.12 ± 0.05. The neutron dose response exhibited a linear relationship (R2=0.9974) within the detectable neutron dose equivalent range under test (0.4-3 mSv). For this neutron field, the lower limit of detection was between 0.2 and 0.4 mSv. Over different post-irradiation storage times of up to 180 days, the readings fluctuated within ±5%. Personal dosimetry based on the investigated OSLN dosimeter is considered to be reliable under similar neutron exposure conditions, i.e. similar neutron energy spectra and dose equivalent values.

  17. Unfolding neutron spectra from simulated response of thermoluminescence dosimeters inside a polyethylene sphere using GRNN neural network

    Science.gov (United States)

    Lotfalizadeh, F.; Faghihi, R.; Bahadorzadeh, B.; Sina, S.

    2017-07-01

    Neutron spectrometry using a single-sphere containing dosimeters has been developed recently, as an effective replacement for Bonner sphere spectrometry. The aim of this study is unfolding the neutron energy spectra using GRNN artificial neural network, from the response of thermoluminescence dosimeters, TLDs, located inside a polyethylene sphere. The spectrometer was simulated using MCNP5. TLD-600 and TLD-700 dosimeters were simulated at different positions in all directions. Then the GRNN was used for neutron spectra prediction, using the TLDs' readings. Comparison of spectra predicted by the network with the real spectra, show that the single-sphere dosimeter is an effective instrument in unfolding neutron spectra.

  18. Neutron spectrum survey around the cyclotron of IEN/Brazilian CNEN: calibration of neutron personnel dosemeter

    International Nuclear Information System (INIS)

    Fajardo, P.W.

    1991-01-01

    The albedo neutron dosimeter is calibrated directly at the work place due to its high energy dependence. This thesis deals with the study, analysis and application of neutron measurement techniques in order to obtain information about the neutron spectrum and neutron dose equivalent at several representative working places of the cyclotron laboratory of the Nuclear Engineering Institute (IEN). These data are employed mainly in the calibration of the brazilian albedo neutron dosimeter. Bonner spheres and foil activation were used in neutron spectra measurements and the neutron dose equivalents were measured with the single sphere albedo technique. BF 3 and 3 He proportional detectors and 6 LiI scintillation detector were also used in these measurements. The single sphere technique turned out to be more appropriate for neutron dosimetry for calibrating the albedo dosimeter in the varying fields of the cyclotron. Calibration the albedo dosimeter in the varying fields of the cyclotron. Calibration factors were found for routine applications, when the workers are protected by shielding and for radiological accident applications, in the case that a worker is exposed inside the cyclotron room. In all situations the performance of the brazilian albedo dosimeter is compared with that of the german albedo dosimeters. (author)

  19. Dosimeter incorporating radiophotoluminescent detectors for thermal neutrons and γ-rays in n-γ fields

    Energy Technology Data Exchange (ETDEWEB)

    Salem, Y.O. [Groupe RaMsEs, Institut Pluridisciplinaire Hubert Curien (IPHC), UMR 7178 CNRS/IN2P3, 23 rue du Loess, BP 28, F-67037 Strasbourg Cedex 2 (France); Nachab, A., E-mail: a.nachab@uca.ma [Département de physique, Faculté Poly-disciplinaire, Université Cadi Ayyad, Route Sidi Bouzid BP 4162, 46000 Safi (Morocco); Roy, C.; Nourreddine, A. [Groupe RaMsEs, Institut Pluridisciplinaire Hubert Curien (IPHC), UMR 7178 CNRS/IN2P3, 23 rue du Loess, BP 28, F-67037 Strasbourg Cedex 2 (France)

    2016-10-15

    We have developed a dosimeter associating different neutron converters with two radiophotoluminescent detectors to measure thermal neutrons and γ-rays in a mixed n-γ field. Tests show that the H{sup ∗}(10) and H{sub p}(10) responses to thermal neutrons and γ-rays are linear with detection limits lower than 0.4 mSv. The angular dependence of the dosimeter response is satisfactory and the influence of a phantom on the results is examined.

  20. Commissioning optically stimulated luminescence in vivo dosimeters for fast neutron therapy

    Energy Technology Data Exchange (ETDEWEB)

    Young, Lori A., E-mail: layoung@uw.edu; Sandison, George [Department of Radiation Oncology, University of Washington, Seattle, Washington 98115 (United States); Yang, Fei [Sylvester comprehensive Cancer Center, University of Miami, Miami, Florida 33124 (United States); Woodworth, Davis [Department of Physics, University of Reno, Reno, Nevada 89557 (United States); McCormick, Zephyr [Department of Physics, University of California, Santa Barbara, California 93106 (United States)

    2016-01-15

    Purpose: Clinical in vivo dosimeters intended for use with photon and electron therapies have not been utilized for fast neutron therapy because they are highly susceptible to neutron damage. The objective of this work was to determine if a commercial optically stimulated luminescence (OSL) in vivo dosimetry system could be adapted for use in fast neutron therapy. Methods: A 50.5 MeV fast neutron beam generated by a clinical neutron therapy cyclotron was used to irradiate carbon doped aluminum oxide (Al{sub 2}O{sub 3}:C) optically simulated luminescence dosimeters (OSLDs) in a solid water phantom under standard calibration conditions, 150 cm SAD, 1.7 cm depth, and 10.3 × 10.0 cm field size. OSLD fading and electron trap depletion studies were performed with the OSLDs irradiated with 20 and 50 cGy and monitored over a 24-h period to determine the optimal time for reading the dosimeters during calibration. Four OSLDs per group were calibrated over a clinical dose range of 0–150 cGy. Results: OSLD measurement uncertainties were lowered to within ±2%–3% of the expected dose by minimizing the effect of transient fading that occurs with neutron irradiation and maintaining individual calibration factors for each dosimeter. Dose dependent luminescence fading extended beyond the manufacturer’s recommended 10 min period for irradiation with photon or electron beams. To minimize OSL variances caused by inconsistent fading among dosimeters, the observed optimal time for reading the OSLDs postirradiation was between 30 and 90 min. No field size, wedge factor, or gantry angle dependencies were observed in the OSLDs irradiated by the studied fast neutron beam. Conclusions: Measurements demonstrated that uncertainties less than ±3% were attainable in OSLDs irradiated with fast neutrons under clinical conditions. Accuracy and precision comparable to clinical OSL measurements observed with photons can be achieved by maintaining individual OSLD calibration factors and

  1. Fast neutron dosimetry using CaSO4:Dy thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Azorin, N.G.; Salvi, C.R.; Rubio, J.L.; Gutierrez, C.A.

    1980-01-01

    The use of CaSO 4 :Dy phosphor powder in fast neutron dose measurements using the activation of sulphur from the 32 S(n,p) 32 P reaction is described. The thermoluminescence induced during the irradiation and that due to the decay of short-lived activation products, is erased by annealing the dosimeters after a post-irradiation time of 3 days. The self-induced thermoluminescence measured at different intervals of post-irradiation time, gives an estimation of the fast neutron dose to which the dosimeters were exposed

  2. Qualification of the monitor Pug-7N like dosimeter for neutrons

    International Nuclear Information System (INIS)

    Benites R, J. L.; Vega C, H. R.; Murillo O, R.; Velazquez F, J. B.

    2011-10-01

    By means of an inter-comparison method, the monitor for neutrons Pug-7N was enabled like dosimeter for neutrons of two magnitudes: the environmental equivalent dose, H*(10), and the H equivalent dose. The monitor Pug-7N has a plastic detector of scintillation Pns-20 that can be used inside or outside of its polyethylene cylindrical moderator. This designed to detect the neutrons presence that is shown in ana logical form by means of a fast count. Although the instrument is useful to detect the neutrons presence its design it does not allow to estimate the dose. With the purpose of enabling it as dosimeter for neutrons, their response was compared with the response of the area monitor for neutrons Bert hold Lb 6411 and Eberline NRD model Asp-1. Under the same irradiation conditions the 3 instruments were exposed to a source of 241 AmBe of 3.7E(9) Bq (100 mCi) of activity whose spectrum and dosimetric magnitudes were determined with a spectrometric system of Bonner spheres with scintillator of 6 Lil(Eu) and the NSDUAZ code. Conversion factors of H*(10)/cpm and H/cpm were obtained for the two options of the monitor detector Pug-7N, with this procedure the monitor Pug-7N besides determining the presence of neutrons, it has been enabled for their use as dosimeter for neutrons. (Author)

  3. Performance testing of personnel extremity dosimeters by Korean LiF: Mg, Cu, Na, Si TLD(KLT-300)

    International Nuclear Information System (INIS)

    Kim, J.L.; Lee, J.I.; Chang, S.Y.; Choi, H.S.; Lee, D.H.; Han, S.J.

    2005-01-01

    Full text: As the needs and opportunities for utilization of atomic energy and radiation are increasing, the related industries, medicines, environments are developing and the relevant organizations and companies are also becoming diverse. In result, the types and kinds of the radiation related to occupational environments are becoming diversified. For the whole body dosimeters, the methodology and criteria for the performance evaluation and safety regulations and laws have been prepared in some detail, but for the extremity dosimeters, those are not prepared yet in Korea. The extremity dosimeters are required when the extremity part of our body, such as hand, elbow, and arm below the elbow, the foot, knee, and leg below the knee are exposed to the radiation in specific work environments. The dosimeter irradiation conditions are clearly discriminated between the whole body exposure condition and the extremity exposure condition. By the investigation and analysis of the management status and dose evaluation methods of the extremity dosimeters for the local absorbed dose, the personnel monitoring system of the extremity dosimeter services in Korea can be diagnosed, and the performance testing criteria and procedures can be established. Therefore, this study presents the performance testing results of extremity dosimeters on the finger and arm/leg phantoms by the procedures recommended in the ANSI (American National Standard) N13.32 using KLT-300 TL materials (LiF:Mg,Cu,Nas,Si) which were developed in Korea Atomic Energy Research Institute (KAERI). The results show that the performance index for the two types of phantoms are sufficiently satisfied with the prescribed tolerance level in the all of the test categories listed in the ANSI N13.32. These results and procedures used in this study can be applicable for regulatory body to establish the standard criteria for acceptable performance and testing conditions for personnel extremity dosimeters services in the

  4. Establishing personal dosimetry procedure using optically stimulated luminescence dosimeters in photon and mixed photon-neutron radiation fields

    International Nuclear Information System (INIS)

    Le Ngoc Thiem; Bui Duc Ky; Trinh Van Giap; Nguyen Huu Quyet; Ho Quang Tuan; Vu Manh Khoi; Chu Vu Long

    2017-01-01

    According to Vietnamese Law on Atomic Energy, personal dosimetry (PD) for radiation workers is required periodically in order to fulfil the national legal requirements on occupational radiation dose management. Since the radiation applications have become popular in Vietnamese society, the thermal luminescence dosimeters (TLDs) have been used as passive dosimeters for occupational monitoring in the nation. Together with the quick increase in radiation applications and the number of personnel working in radiation fields, the Optically Stimulated Luminescence Dosimeters (OSLDs) have been first introduced since 2015. This work presents the establishment of PD measuring procedure using OSLDs which are used for measuring photons and betas known as Inlight model 2 OSL (OSLDs-p,e) and for measuring mixed radiations of neutrons, photons and betas known as Inlight LDR model 2 (OSLDs-n,p,e). Such following features of OSLDs are investigated: detection limit, energy response, linearity, reproducibility, angular dependency and fading with both types of OSLDs-p,e and OSLDs-n,p,e. The result of an intercomparison in PD using OSLDs is also presented in the work. The research work also indicates that OSL dosimetry can be an alternative method applied in PD and possibly become one of the most popular personal dosimetry method in the future. (author)

  5. Radiation detectors for personnel monitoring - current developments and future trends

    International Nuclear Information System (INIS)

    Kannan, S.

    2003-01-01

    The radiation detectors for personnel monitoring range from the conventional passive dosimeters like the film badge and the TLD, to sophisticated active dosimeters for integrated gamma, beta and neutron dose measurement. With the availability of high accuracy active dosimeters, the process of personnel monitoring, acceptability among radiation workers, record keeping and dose control have become more simplified. However the high level of sophistication in the active dosimeter has its own inevitable price tag and the new breed of active dosimeters are prohibitively costly. The silver lining, in the otherwise dark cost scenario of these dosimeters is the potential for cost reduction at least in some of the dosimeters in the near future

  6. Development of a semiconductor neutron dosimeter with a PIN diode

    International Nuclear Information System (INIS)

    Kim, Seungho; Lee, Namho; Cho, Jaiwan; Youk, Geunuck

    2004-01-01

    When a Si PIN diode is exposed to fast neutrons, it produces displacement in Si lattice structure of the diode. Defects induced from structural dislocation become effective recombination centers for carriers which pass through the base of a PIN diode. Hence, increasing the resistivity of the diode decreases the current for the applied forward voltage. This paper involves the development of a neutron sensor based on the phenomena of the displacement effect damaged by neutron exposure. The neutron effect on the semiconductor was analyzed, and multi PIN diode arrays with various intrinsic layer (I layer) thicknesses and cross sections were fabricated. Under irradiation tests with a neutron beam, the manufactured diodes have good characteristics of linearity in a neutron irradiation experiment and give results that the increase of thickness of I layer and the decrease of the cross-section of the PIN diodes improve the sensitivity. Newly developed PIN diodes with a thicker I layer and various cross sections were retested and showed the best neutron sensitivity in the condition that the I layer thickness was similar to the length of a side of the cross-section. On the basis of two test results, final PIN diodes with a rectangular shape were manufactured and the characteristics for neutron detectors were analyzed through the neutron beam test using the on-line electronic dosimetry system. The developed PIN diode shows a good linearity to absorbed dose in the range of 0 to 1,000cGy (Tissue) and its neutron sensitivity is 13 mV/cGy at a constant current of 5 mA, that is three higher than that of similar commercially developed neutron detectors. Moreover the device shows less dependency on the orientation of the neutron beam and a considerable stability in an annealing test for a long period. (author)

  7. CMOS-Based Neutron Spectroscopic Dosimeter (CNSD), Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — Monitoring space radiation and the dose received by astronauts is important, especially for future long-duration missions. Neutrons contribute a significant...

  8. Response of CR-39 based personnel neutron dosemeter in terms of directional dose equivalent, in free air and on phantom

    International Nuclear Information System (INIS)

    Pal, Rupali R.; Sathian, Deepa; Jayalakshmi, V.; Chougaonkar, M.P.

    2011-01-01

    CR-39 is the most sensitive of nuclear track detectors for protons and is recommended as an effective neutron dosimeter because of it's low threshold energy of 100 keV neutrons. The fraction of protons that gives detectable tracks in CR-39 depends on the energy of the proton angle of incidence and etching conditions. As a consequence the registration efficiency of neutrons in the CR-39 plastics used for neutron personnel monitoring is strongly influenced by the direction of radiation incidence. This paper presents the relative response of CR-39 at varying neutron incident angles, for 241 Am-Be neutron source spectra in free air and on ISO phantom, in terms of operational quantities. It is observed that the angular dependence of CR-39 for irradiations in air and on phantom is essentially the same indicating that the phantom does not affect the directional response of CR-39. (author)

  9. Design and development of wide energy neutron REM equivalent spectrometer-dosimeters based on polycarbonates and Cr-39

    International Nuclear Information System (INIS)

    Faermann, S.

    1985-03-01

    This work describes a system composed of a Rem response personnel neutron dosemeter, based on boron radiators and a polycarbonate track detector, for monitoring dose equivalents in the energy range 1 eV to 14 MeV, an electrochemical etching system for revealing damage sites in solid state track etch detectors, a reader for magnifying the etched pits and a microprocessor for evaluating the dose equivalents and their uncertainties. The performance and directional dependence of the dosemeter when exposed to monoenergetic and polyenergetic neutron fields in the epithermal and fast energy regions are discussed. Saturation effects in polycarbonate foils are presented and a comparison is made between the response of polycarbonate and CR-39 foils, used as passive detectors in the dosemeter. A new passive miniature fast neutron spectrometer-dosimeter is also described. The device is based on the detection of proton tracks by electrochemical etching of CR-39 foils covered with thin polyethylene layers of different thicknesses. By means of this device it is possible to assess the fast neutron energy spectrum in 10 energy intervals in the energy range 0.5-15 MeV. Dose equivalents can be determined in the dose equivalent range 20 mRem to 8 Rem, approximately (author)

  10. Results of the systematic study of neutron dosimeters and neutron radiometers responses from Bruyeres and Valduc's experiments

    International Nuclear Information System (INIS)

    Pras, Ph.; Ledoux, X.; Patin, Y.

    1999-01-01

    This document gives the results of the systematic study of neutron dosimeters (Bubbles detectors) and neutron radiometers (Cramal, Nausicaa, EGG Lb6411) with standard sources and with the Bruyeres Van de Graaff 4MV accelerator. The dose equivalent rate response as a function of the neutron energy is parameterized. Even for low dose equivalent rate, a good reproducibility of the measurements is found in the strict respect of a given method. The response of the different systems is independent of the dose equivalent rate. (author)

  11. Fast Neutron Dosimeter for the Space Environment, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Model calculations and risk assessment estimates indicate that secondary neutrons, with energies ranging between 0.5 to >150 MeV, make a significant contribution...

  12. Calibration and automatic counting in a neutrons dosimeter

    International Nuclear Information System (INIS)

    Tavera-Davila, M.L.

    1991-01-01

    Over the past decade, the majority of improvements in spectrometry have not come from improved detectors but from better computing facilities, in Folding codes, and nuclear data. However several types of passive neutron detectors have the potential for development into crude, low resolution spectrometers, to be worn by individuals, making them attractive to be subject of more time of investigation. This thesis contributes to the understanding of the properties of solid state nuclear tracks (SSNTD) in order to use them in neutron dosimetry and spectrometry. In the first chapter the basic principles on neutron interaction with matter and some relevant considerations about detection and dosimetry are presented. The third chapter deals with detection properties of solid state nuclear track detectors (SSNTD). Calibration methods are presented for three different applications, detection of heavy ions, radon and neutrons. In chapter 4, some equations are derived which predict the whole efficiency formation of thin plastic detectors as a function of range, angle, and type of inciding particle as well as residual thickness. Using experimental data on proton registration properties in thin plastic detectors and the former derived equations, mathematical expressions are developed to predict the sparking counting efficiency for recoll protons and trions produced by neutrons in SSNTD. Chapter five deals with the comparison between experimental results and theoretical results of chapter 4. In chapter 6 experimental optimum parameters for electrochemical etching of thin plastic detectors are presented. The electrochemical etching efficiency is compared with the spark countering efficiency obtained in chapter 5. In chapter 7, general comments on applications to neutron dosimetry are presented. (Author)

  13. Photon and neutron energy response of Thermoluminescent (TL) dosimeters

    International Nuclear Information System (INIS)

    Thilagam, L.; Priya, M.R.; Mohapatra, D.K.

    2018-01-01

    Theoretical Monte Carlo (MC) simulations are carried out to investigate the relative thermoluminesence (TL) response of the most commonly used TLD materials to a wide range of photon energy. The effect of polytetrafluoroethylene (PTFE) on TL response of CaSO 4 :Dy is also studied. Additionally, the neutron response of LiF:Mg,Ti TL materials with different concentrations of 6 Li is estimated in terms of the number of 6 Li(n, t) 4 He capture reactions for a wider neutron energy

  14. MONTE CARLO CALCULATION OF THE ENERGY RESPONSE OF THE NARF HURST-TYPE FAST- NEUTRON DOSIMETER

    Energy Technology Data Exchange (ETDEWEB)

    De Vries, T. W.

    1963-06-15

    The response function for the fast-neutron dosimeter was calculated by the Monte Carlo technique (Code K-52) and compared with a calculation based on the Bragg-Gray principle. The energy deposition spectra so obtained show that the response spectra become softer with increased incident neutron energy ahove 3 Mev. The K-52 calculated total res nu onse is more nearly constant with energy than the BraggGray response. The former increases 70 percent from 1 Mev to 14 Mev while the latter increases 135 percent over this energy range. (auth)

  15. The application of thermoluminescent dosimeters to the measurement of thermal neutron distributions in bulk media

    International Nuclear Information System (INIS)

    Bubb, I.F.; Tang, J.C.N.

    1981-01-01

    The work described was designed to investigate the suitability of measuring neutron flux distributions in bulk media using neutron sensitive thermoluminescent dosimeters, with the initial interest being in wet coal matrices. The phosphors used were 6 LiF and 7 LiF 3.2mmx3.2mmx0.9mm chips. As 7 LiF is sensitive to gamma rays only it was used to correct for the gamma-ray response of 6 LiF

  16. Tenth DOE workshop on personnel neutron dosimetry

    International Nuclear Information System (INIS)

    1984-06-01

    The purpose of this workshop is to promote the international exchange of information on neutron dosimetry. The development of an accurate real-time dosemeter is an immediate need which must be met. Assessment of the neutron dose equivalent at low doses with a reasonable degree of accuracy must be accomplished to provide validity to exposure records. These and other aspects of personal neutron dosimetry are discussed. Separate abstracts have been prepared for each paper for inclusion in the Energy Data Base

  17. Current situation for exoelectron dosimeters of BeO ceramic in neutron dosimetry

    International Nuclear Information System (INIS)

    Gammage, R.B.

    1977-01-01

    Much of the early enthusiasm for using exoelectron dosimeters (ceramic BeO Thermalox 995) in neutron dosimetry was predicted on the belief that the response to fast neutrons, relative to gamma rays, was 0.18 to 0.28 on a R/sub γ/ equiv/tissue rad n/sub f/ basis for neutron energies between 0.1 and 16 MeV. Pairs of BeO disks had to be used, one covered with a polyethylene radiator for producing recoil protons, and the other covered with Teflon. More recent studies indicated a considerably lower ratio of 0.11 for Health Physics Reactor Research fission neutrons. In the earlier work the BeO was coated with gold to enhance the surface conductivity during reading of the thermally stimulated exoelectron emission (TSEE). No metallic coating is now deemed to be necessary. Perhaps thermal neutron contamination of the fast neutron beams due to some thermalization within the hydrogenous radiator was sufficient to cause the high apparent fast neutron sensitivity via n, γ reactions. Whatever the cause, however, the lower value of 0.11 has caused a marked subsidence of enthusiasm in this technique of fast neutron monitoring

  18. Development of direct reading dosimeters for the dose 0-3 mSv and 0-5 mSv ranges for personnel monitoring

    International Nuclear Information System (INIS)

    Gaikwad, P.V.; Shirkar, Y.B.; Patil, A.S.; Madgaonkar, P.P.; Kale, K.L.; Guhagarkar, H.V.; Gandhi, D.P.; Gupta, S.K.; Kothiyal, G.P.; Sahni, V.C.

    1998-01-01

    Direct reading dosimeters (DRDs) are widely used to measure cumulative dose received by personnel working at nuclear reactor sites or in other environment having x- and gamma rays. A DRD operates on the principle of gold leaf electroscope, and is a small, rugged, hermetically sealed, self reading type device easily carried by an individual in his pocket. The development of dosimeters suitable for the dose ranges 0-3 mSv and 0-5 mSv is reported

  19. New developments of Czech Personnel Neutron Dosemeter and its application

    International Nuclear Information System (INIS)

    Trousil, J.; Spurny, F.; Kralik, M.; Zelenka, Z.

    2001-01-01

    There are two systems for neutron personal dosimetry used at practice in the Czech Republic: - solid state nuclear track detectors in contact with fissionable materials (intermetallic enriched U + AI alloy, and 232 Th) in Cd shield badge, and - Albedo dosimeter based on 6 LiF and 7 LiF thermoluminescent detectors in a plastic badge covered from the front side by Cd to screen it from directly coming thermalized neutrons. The first system is used in National Personal Dosimetry Service; its performance is continuously checked and improved. In this contribution we would like to present some of results of this development, the contribution to the calibration of the second system included. (authors)

  20. Neutron skyshine measurement at a K1200 superconducting heavy ion cyclotron using bubble dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Mukherjee, B. [Safety Div., Australian Nuclear Science and Technology Organisation, Menai (Australia); Ronningen, R.M. [Michigan State Univ., National Superconducting Cyclotron Lab., East Lansing, MI (United States); Rossi, P. [Michigan State Univ., Office of Radiation, Chemical and Biological Safety, East Lansing, MI (United States)

    1999-07-01

    Understanding the characteristics of the neutron skyshine radiation is necessary for an accurate assessment of the environmental dose in the vicinity of the containment of a high-energy particle accelerator. At the National Superconducting Cyclotron Laboratory (NSCL), neutron skyshine was measured, using beams of 140 MeV/nucleon {sup 4}He and 80 MeV/nucleon {sup 22}Ne ions from the K1200 superconducting cyclotron. After passing through a radioactive-beam production target, the ion beam stopped in a solid aluminium stopping bar inside of a dipole magnet, resulting in the production of high energy fragmentation as well as evaporation neutrons in the NSCL Analysis Hall. The neutron dose equivalent and energy spectrum at the 1.37 m thick concrete roof of the Analysis Hall, directly above the aluminium target bar (reference point), were estimated, using a spherical 'rem-counter' and a set of seven Bonner-spheres, respectively. The skyshine dose, from neutrons transmitted through 21.5-cm local iron 'shielding' of the dipole magnet and the concrete roof, were evaluated using superheated bubble dosimeters at 50 m, 75 m, 100 m and 115 m from the reference point. The neutron doses beyond the extremity of the NSCL facility were extrapolated from the results of this investigation and were used to predict the exposure to members of the public by considering the operation schedule of the K1200 cyclotron. (authors)

  1. SU-E-T-75: Commissioning Optically Stimulated Luminescence Dosimeters for Fast Neutron Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Young, L [UniversityWashington, Seattle, WA (United States); Yang, F; Sandison, G [University of Washington, Seattle, WA (United States); Woodworth, D [University of California, Santa Barbara, Santa Barbara, CA (United States); McCormick, Z [University of Nevada - Reno, Reno, Nevada (United States)

    2014-06-01

    Purpose: Fast neutrons therapy used at the University of Washington is clinically proven to be more effective than photon therapy in treating salivary gland and other cancers. A nanodot optically stimulated luminescence (OSL) system was chosen to be commissioned for patient in vivo dosimetry for neutron therapy. The OSL-based radiation detectors are not susceptible to radiation damage caused by neutrons compared to diodes or MOSFET systems. Methods: An In-Light microStar OSL system was commissioned for in vivo use by radiating Landauer nanodots with neutrons generated from 50.0 MeV protons accelerated onto a beryllium target. The OSLs were calibrated the depth of maximum dose in solid water localized to 150 cm SAD isocenter in a 10.3 cm square field. Linearity was tested over a typical clinical dose fractionation range i.e. 0 to 150 neutron-cGy. Correction factors for transient signal fading, trap depletion, gantry angle, field size, and wedge factor dependencies were also evaluated. The OSLs were photo-bleached between radiations using a tungsten-halogen lamp. Results: Landauer sensitivity factors published for each nanodot are valid for measuring photon and electron doses but do not apply for neutron irradiation. Individually calculated nanodot calibration factors exhibited a 2–5% improvement over calibration factors computed by the microStar InLight software. Transient fading effects had a significant impact on neutron dose reading accuracy compared to photon and electron in vivo dosimetry. Greater accuracy can be achieved by calibrating and reading each dosimeter within 1–2 hours after irradiation. No additional OSL correction factors were needed for field size, gantry angle, or wedge factors in solid water phantom measurements. Conclusion: OSL detectors are a useful for neutron beam in vivo dosimetry verification. Dosimetric accuracy comparable to conventional diode systems can be achieved. Accounting for transient fading effects during the neutron beam

  2. Development of neutron personnel monitoring system based on CR-39 solid state nuclear track detector

    International Nuclear Information System (INIS)

    Massand, O.P.; Kundu, H.K.; Marathe, P.K.; Supe, S.J.

    1990-01-01

    Personnel neutron monitoring aims at providing a method to evaluate the magnitude of the detrimental effects on the personnel exposed to neutrons. Neutron monitoring is done for a small though growing number of personnel working with neutrons in a wide range of situations. Over the years, many solid state nuclear track detectors (SSNTD) have been tried for neutron personnel monitoring. CR-39 SSNTD is a proton sensitive polymer and offers a lot of promise for neutron personnel monitoring due to its high sensitivity and lower energy threshold for neutron detection. This report presents the mechanism of track formation in this polymer, the development of this neutron personnel monitoring system in our laboratory, its various characteristics and its promise as a routine personnel neutron monitor. (author). 1 tab., 7 figs

  3. An attempt to fabricate an ion injection type fast neutron dosimeter

    International Nuclear Information System (INIS)

    Pells, G.P.; Hughes, A.E.

    1976-10-01

    This report describes the results of work carried out to investigate a novel type of solid state, fast neutron dosimeter. The concept and quantitative features of the dosimeter, in which collisions between fast neutrons and ions in a source compound cause the ions to be injected into a host compound where they could then be detected by their characteristic luminescence, are described. It is shown that with the aid of a complex reader mrad sensitivities should in principle be possible. Initial experiments showed that accelerator implantation of Bi 3+ ions into CaO gives a linear dependence of luminescence output with fluence of activator ions. A number of host/activator combinations were investigated from which SiO 2 :Gd 3+ appeared the most promising. Host/activator configurations are briefly considered. The production of compacts of SiO 2 and Gd 2 O 3 ultrafine powders is described. Attempts were made to characterise these compacts using scanning electron microscopy and luminescence spectroscopy. After fast neutron irradiation to 250 Mrad no Gd 3+ ion injection could be detected. An alternative configuration of ultrafine SiO 2 powder dispersed in a solid Gd(NO 3 ) 3 .5H 2 O matrix is described but again no Gd 3+ ion injection into the SiO 2 could be detected after fast neutron irradiation. Perspex, as a host, was also tried without success. An appendix is included which demonstrates that the lack of success in detecting ion-injection may be largely due to the optically opaque nature of the specimens that could be produced. (author)

  4. Application of semiconductor MOSFET and pin diode dosimeters to epithermal neutron beam dose distribution measurements in phantoms

    International Nuclear Information System (INIS)

    Carolan, M.G.; Wallace, S.A.; Allen, B.J.; Rosenfeld, A.B.; Mathur, J.N.

    1996-01-01

    For any clinical application of Boron Neutron Capture Therapy (BNCT) fast and accurate dose calculations will be required for treatment planning. Such calculations are also necessary for the planning and interpretation of results from pre-clinical and clinical trials where the speed of calculation is not so critical. A dose calculation system based on the MCNP Monte Carlo Neutron transport code has been developed by Wallace. This system takes image data from CT scans and constructs a voxel based geometrical model for input into MCNP. To validate the calculations, a number of phantoms were constructed and exposed in the HB11 epithermal neutron beam at the HFR of the CEC Joint Research Centre in Petten. The doses recorded by arrays of PIN diode neutron dosimeters and MOSFET gamma dosimeters in these phantoms were compared with the calculated results from the MCNP dose planning system. Initial results have been reported elsewhere. Poster 197. (author)

  5. Qualification of the monitor Pug-7N like dosimeter for neutrons; Habilitacion del monitor PUG-7N como dosimetro para neutrones

    Energy Technology Data Exchange (ETDEWEB)

    Benites R, J. L. [Centro Estatal de Cancerologia de Nayarit, Av. Enfermeria, Fracc. Fray Junipero Serra, 63000 Tepic, Nayarit (Mexico); Vega C, H. R.; Murillo O, R. [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Calle Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico); Velazquez F, J. B., E-mail: jlbenitesr@prodigy.net.mx [Universidad Autonoma de Nayarit, Postgrado CBAP, Carretera Tepic Compostela Km. 9, Xalisco, Nayarit (Mexico)

    2011-10-15

    By means of an inter-comparison method, the monitor for neutrons Pug-7N was enabled like dosimeter for neutrons of two magnitudes: the environmental equivalent dose, H*(10), and the H equivalent dose. The monitor Pug-7N has a plastic detector of scintillation Pns-20 that can be used inside or outside of its polyethylene cylindrical moderator. This designed to detect the neutrons presence that is shown in ana logical form by means of a fast count. Although the instrument is useful to detect the neutrons presence its design it does not allow to estimate the dose. With the purpose of enabling it as dosimeter for neutrons, their response was compared with the response of the area monitor for neutrons Bert hold Lb 6411 and Eberline NRD model Asp-1. Under the same irradiation conditions the 3 instruments were exposed to a source of {sup 241}AmBe of 3.7E(9) Bq (100 mCi) of activity whose spectrum and dosimetric magnitudes were determined with a spectrometric system of Bonner spheres with scintillator of {sup 6}Lil(Eu) and the NSDUAZ code. Conversion factors of H*(10)/cpm and H/cpm were obtained for the two options of the monitor detector Pug-7N, with this procedure the monitor Pug-7N besides determining the presence of neutrons, it has been enabled for their use as dosimeter for neutrons. (Author)

  6. Multisphere system neutron spectrometry applied to dosimetry for the personnel

    International Nuclear Information System (INIS)

    Allinei, P.G.

    1992-01-01

    Neutron dosimetry is a necessity that must be dealt with in order to ensure efficient monitoring of all personnel regarding radiology safety. Dosimetric variables are difficult to measure for they are dependent on complex functions evolving with the energy of neutrons, which forces us to determine their energetic distribution. We have chosen to use the multisphere system associated to an unfolding code in order to perform neutron spectrometry, our purpose being to determine these dosimetric variables. The initial stage consists in modifying a research code, the code SOHO, in order to adapt it to our needs. The resulting new version was subsequently tested and proven successful by means of computerized simulations. Afterwards, we used reference dosimetric and spectral beams to confirm the position results previously obtained. At the time of this test, the code SOHO yielded results coherent with the theoretical values, and even allowed the quantity of radiation diffused by the laboratory structures to be estimated. The final part of this study consists in applying the previously perfected technique to authentic situations. The results thus obtained are compared to those obtained by conventional methods in order to reveal the interest of neutron spectrometry used for dosimetry of the personnel

  7. Effects of additives and impurities on the properties of CR-39 neutron dosimeters

    International Nuclear Information System (INIS)

    Tarle, G.

    1982-01-01

    Several problems have plagued users of commercial CR-39 for neutron dosimetry work. Large variations in sensitivity have been observed in samples received from different manufacturers, from different batches from the same manufacturer and even within a given sheet. The latter variations have taken several forms. Clear etching regions of lower sensitivity near the gasket and sometimes extending well away from the gasket in drip-like patterns have been observed. Large demarcated patches of differing sensitivity and sensitivity variations with depth have been found in many commercial samples of CR-39. In addition, we have observed a general roughening of the post-etch surface and a patchiness of bulk etching associated with contact patterns of adjacent sheets, which have been stored for extended (greater than or equal to 6 mos) periods of time. All of these problems combine to make commercially available CR-39 unreliable as a neutron dosimeter. In an attempt to identify and correct the causes of these problems, a systematic study of the effects of additives, impurities and manufacturing conditions on the uniformity of CR-39 track detectors has been undertaken, and is reported in this paper

  8. ON THE ADVISABILITY OF REJECTION OF SUBTRACTION OF BACKGROUND DOSIMETERS INDICATION VALUES FROM THE MEASURED VALUES OF INDIVIDUAL DOSES OF OCCUPATIONAL EXPOSURE OF A-GROUP PERSONNEL

    Directory of Open Access Journals (Sweden)

    A. N. Barkovskiy

    2013-01-01

    Full Text Available An article contains substantiation of advisability of rejection of subtraction of background dosimeters indication values from the measured values of individual doses of occupational exposure of A-group personnel in the process of carrying out of individual dosimetry control of medical organizations stuff which was implemented in the Methodical Guideline 2.6.1.3015-12. It is shown that mentioned procedure is firstly incorrect when we deal with low doses, when the impact of natural exposure could be really considerable and secondary senseless when we deal with dose values close to the dose limit value for A-group personnel when the impact of natural exposure is much lower then the error of measurements. It is suggested to carry out the account of the impact of natural exposure into personnel average doses basing on the analysis of statistical distributions of personnel individual doses and on indications of background dosimeters. It is suggested to add the information about indications of background dosimeters into the Base of Federal Statistical Supervision № 1-DOS as it is prescribed in the Methodical Guideline 2.6.1.3015-12. This article is the respond to the article of A.I. Grigoriev published in the present issue of Journal, in which the author contests the reasonableness of this provision.

  9. Stable Chemical Dosimeters for Partial Reconstruction of Nuclear Accident Conditions

    Energy Technology Data Exchange (ETDEWEB)

    Dvornik, I.; Zec, U.; Baric, M.; Razem, D. [Ruder Boskovic Nuclear Institute, Zagreb, Yugoslavia (Croatia)

    1969-10-15

    The application of chemical dosimeters, tissue equivalent with respect to gamma rays and neutrons, is proposed for dosimetric topography of the space around nuclear devices in case of accidents. The dosimeters in the form of sealed glass ampoules have sufficient sensitivity and long-term stability and are evaluated or checked directly by conventional spectrophotometry. The sensitivity, expressed as yield per rad, is approximately equal for gamma rays and neutrons. The resolution in both cases is about one rad, and the range is up to several thousand rads. The precision of dosimetry is {+-} 1 rad or {+-} 2%, whichever is higher. In free space and unshielded the dosimeter measures the total rad-absorbed dose delivered by gamma rays and neutrons, i.e. the first collision gamma plus neutron dose. If used on- or in-phantom, especially if several dosimeters are disposed within and around the same phantom, it can give important data about the amount of the neutron component of the dose and about the effective mean energy of incident neutrons. The neutron component of the dose can be directly measured if the gamma dosimeter is used together with the chemical dosimeter. The experiments giving the change of optical density per rad and the radiation chemical yield with respect to the absorbed dose delivered by 14-MeV neutrons are described in detail. The possibility is also mentioned of applying the dosimeter as a very sensitive monitor for thermal neutrons, which is due to the chlorine content of 4.73% and activation to {sup 38}Cl. The opinion is expressed that this dosimeter deserves some attention as a part of future planning and development work on area and personnel accidental dosimetry systems. (author)

  10. Evaluation of the implementation and use of active personal dosimeters for neutrons in Brazil; Evaluacion de la implementacion y del uso de los dosimetros individuales activos para neutrones en el Brasil

    Energy Technology Data Exchange (ETDEWEB)

    Castro B, C. P.; Wagner P, W.; De Souza P, K. C., E-mail: karla@ird.gov.br [Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende s/n, Recreio Bandeirantes, 22780-160 Rio de Janeiro (Brazil)

    2014-08-15

    This work was conducted through of a field research based on a questionnaire sent to users of active personal dosimeters. A retrospective study of the last six years was also carried out of the services in the Neutron Metrology Laboratory (2008-2013) referent to the active personal dosimeters, taking into consideration the standards ISO-8529-3 and IEC-61526. The active personal dosimeters are defined as any instrument of individual monitoring with direct reading capacity, used by individuals exposed to ionizing radiation fields. Through research was verified that the active personal dosimeters work associated with other dosimeter types. Considering all dosimeters declared in the questionnaire, only two dosimeters (MGP brand Dmc 2000-GN model and the brand ATOMTEX model AT2503A) have conformity declaration with the international standard IEC-61526: 2005 reported by the manufacturers. (author)

  11. Chemical methods for the use of niobium from pressure vessel cladding as a fast neutron dosimeter

    International Nuclear Information System (INIS)

    Karnani, Hari

    1986-08-01

    the steel samples from the cladding of a pressure vessel of an operating nuclear power reactor were obtained by scraping. The cladding material of the pressure vessel contained about 0.5 % niobium. It was desired to use the niobium as a dosimeter for estimating fast fluences at the pressure vessel. The weak radiation from the reaction product 93m Nb cannot be measured in the presence of other elements and interfering activities. A method was developed to separate niobium from other metals present; the concentration and yield of niobium were determined spectrophotometrically. The irradiated niobium was electrodeposited from aqueous solutions on copper discs. The amount of the deposited niobium was determined by a radiochemical method which makes use of its own radioactivity - measured with a liquid scintillation counter - and the known starting mass of niobium. It was possible to determine the deposited niobium masses (5 to 200 microgram) with a desired degree of accuracy. The absolute emission rate of X-rays could then be measured without any self-absorption or interference from other activities. The mass of niobium on each preparate and its X-ray emission rate, later on, were used as basic experimental data for the estimation of last neutron doses at the pressure vessel

  12. Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations.

    Science.gov (United States)

    Nedaie, Hassan Ali; Darestani, Hoda; Banaee, Nooshin; Shagholi, Negin; Mohammadi, Kheirollah; Shahvar, Arjang; Bayat, Esmaeel

    2014-01-01

    High-energy linacs produce secondary particles such as neutrons (photoneutron production). The neutrons have the important role during treatment with high energy photons in terms of protection and dose escalation. In this work, neutron dose equivalents of 18 MV Varian and Elekta accelerators are measured by thermoluminescent dosimeter (TLD) 600 and TLD700 detectors and compared with the Monte Carlo calculations. For neutron and photon dose discrimination, first TLDs were calibrated separately by gamma and neutron doses. Gamma calibration was carried out in two procedures; by standard 60Co source and by 18 MV linac photon beam. For neutron calibration by (241)Am-Be source, irradiations were performed in several different time intervals. The Varian and Elekta linac heads and the phantom were simulated by the MCNPX code (v. 2.5). Neutron dose equivalent was calculated in the central axis, on the phantom surface and depths of 1, 2, 3.3, 4, 5, and 6 cm. The maximum photoneutron dose equivalents which calculated by the MCNPX code were 7.06 and 2.37 mSv.Gy(-1) for Varian and Elekta accelerators, respectively, in comparison with 50 and 44 mSv.Gy(-1) achieved by TLDs. All the results showed more photoneutron production in Varian accelerator compared to Elekta. According to the results, it seems that TLD600 and TLD700 pairs are not suitable dosimeters for neutron dosimetry inside the linac field due to high photon flux, while MCNPX code is an appropriate alternative for studying photoneutron production.

  13. Dosimetric evaluation of the Fricke gel dosimeter using the spectrophotometric technique for application in electron and neutron dosimetry

    International Nuclear Information System (INIS)

    Mangueira, Thyago Fressatti

    2009-01-01

    In this work the main dosimetric characteristics of the Fricke Xylenol Gel (FXG) solution were established for further application in the measurement of dose distribution of clinical electron fields. The dose-response curves of the FXG in a neutron field were also evaluated for the research in Boron Neutron Capture Therapy (BNCT) and industrial electron fields. The standard reading technique was the spectrophotometric. For the clinical field, the intra and inter-batch reproducibility are better than 1.4% and 5.1 %, respectively, the response presents a linear behavior for doses ranging from 0.2 to 40 Gy independently of the energy and the dose rate in the studied ranges. Due to the effects of the FXG natural oxidation, the optimum elapsed time between FXG preparation and irradiation was established as 24h period and the behavior of the dose-response curve of the FXG using the variation in the absorbance relative to the non-irradiated dosimeter as a basis during the whole studied period were not altered. The dose-response to the industrial electron beam presented an exponential decreasing behavior and the neutron beam for research in BNCT presented a linear behavior for the complete studied dose range. According to the obtained results for the different types of radiation studied for the FXG, there was no change in the position of the characteristic bands of the absorption spectrum due to the interaction of these radiation types. Additional tests were performed to determine the digital photographic imaging of FXG analyses viability and the application of FXG dosimetry on intracavitary brachytherapy. The good performance of the FXG dosimeter in the tests that were carried out indicates that this dosimeter may be applied to the tri-dimensional dose evaluation in radiotherapic treatments using electrons and neutron beams. (author)

  14. Evaluation of neutron dosimetry techniques for well-logging operations

    International Nuclear Information System (INIS)

    Cummings, F.M.; Haggard, D.L.; Endres, G.W.R.

    1985-07-01

    Neutron dose and energy spectral measurements from 241 AmBe and a 14 MeV neutron generator were performed at a well-logging laboratory. The measurement technique included the tissue equivalent proportional counter, multisphere, two types of remmeters and five types of personnel neutron dosimeters. Several source configurations were used to attempt to relate data to field situations. The results of the measurements indicated that the thermoluminescent albedo dosimeter was the most appropriate personnel neutron dosimeter, and that the most appropriate calibration source would be the source normally employed in the field with the calibration source being used in the unmoderated configuration. 7 refs., 35 figs., 14 tabs

  15. Evaluation of conditions of radiation protection of medical personnel in intracavitary neutron therapy of cervical cancer

    International Nuclear Information System (INIS)

    Kostromina, K.N.; Korenkov, I.P.; Bocharov, A.L.; Gladkikh, N.N.

    1991-01-01

    Combined radiation therapy was provided to cervical cancer patients. Working conditions of personnel were examined, the rate of exposure doses and flows of neutrons at working places were measured, dose exposures of the personnel were evaluated. It has been concluded that occupational conditions for the medical personnel are considered to be relatively safe

  16. Practical consequences for the use of a personal dosimeter for fast neutrons based on CR39 exposed up to one year

    International Nuclear Information System (INIS)

    Boschung, Markus; Fiechtner, Annette; Mayer, Sabine; Wernli, Christian

    2008-01-01

    Full text: At the Paul Scherrer Institut a personal neutron dosimetry system based on chemically etched CR-39 detectors and automatic track counting is in routine use since 1998. In its original design, the dosimeter is sensitive to thermal neutrons and to neutrons in the energy range from 200 keV up to several MeV. The standard exposition period is 3 months. Recently, a novel concept for individual monitoring was implemented at CERN. In this concept, each worker who possibly enters a radiation zone is equipped with a combined dosimeter for the measurement of personal photon and neutron doses. The dosimeter for photon dose measurement has an instant readout capability and dose measurements are done monthly. The dosimeter for neutron measurement is based on CR-39 detectors and is sensitive to fast neutrons only. The CR-39 detector is only evaluated and a neutron dose determined if the monthly personal photon dose exceeds 2 mSv or if the exposition period of the neutron dosimeter exceeds one year. This novel regime of use of the neutron dosimeter has had some important consequences for its practical implementation. A priori, the wearing period of a neutron dosimeter is not known and can range from 1 month up to 12 or even more months. A good knowledge of the long-term behaviour and characteristics of the detector material is needed. But also organisational and administrative issues have to be considered. The paper will outline the adopted procedure covering not only technical but also organisational aspects. The long-term behaviour of background track density and response to 241 Am-Be over one year are described as well as calibrations performed with 241 Am-Be and 252 Cf sources and in the High-Energy Reference Field Facility at CERN (CERF). The concept of individual monitoring at CERN could be transferred to other locations with high energy accelerators such as PSI and DESY. The experience gained with the neutron dosimeter based on CR-39 since introduction of the

  17. Use of cellulose nitrate plastic track detectors in neutron personnel monitoring

    International Nuclear Information System (INIS)

    Venkataraman, G.; Marathe, P.K.; Joshi, R.V.

    1975-01-01

    Cellulose nitrate, which is a sensitive plastic material wherein even proton tracks could be recorded, was studied with a view to using it for personnel neutron monitoring work. It was found that among the commercially available plastics, the colourless transparent Daicel 6000 variety having a thickness of 0.6mm is satisfactory form the point of view of track recognition. The material was exposed to thermal neutrons, fission neutrons, neutrons from an Am-Be source and to 14 MeV neutrons. As is to be expected, there is no thermal neutron sensitivity. The sensitivity to fast neutrons is less by a factor of two as compared to that of the NTA nuclear track emulsion. It was observed that the background tracks found in the plastic correspond to nearly 100 mrem of fast neutron dose equivalent. It is felt that at present cellulose nitrate is useful for monitoring personnel involved in nuclear criticality accidents. (author)

  18. The dielectric track and thermoluminescent detectors applied to neutron dosimetry in personnel monitoring

    International Nuclear Information System (INIS)

    Mebhah, D.

    1984-03-01

    The personnal dosimeter for neutron based on the detection of fission fragments from 237 Np and 232 Th by a polycarbonate 10 gm, and lithium fluorite 6 LIF/ 7 LIF, allow to cover an energy spectrum from 0.05 eV to 14 MeV with a easy neutron gamma discrimination. In criticality dosimetry, the energy spectrum of the incident neutrons can be defined by two components: the fast component by E b exp(-ae) with E between 0.1 and 14 MeV, a and b determined by a combination of 237 Np and 232 Th track detector responses, and the epithermal component in 1/E, the thermal component having a minor contribution to the total equivalent dose. We took into account the body influence on the detectors response by introducing effective cross section. The equivalent dose obtained by this dosimeter is 20% overestimated in low doses dosimetry. The interpretation of the detectors responses is based on the definition of a factor and a calibration parameter for each zone in which the spectrum is constant. The knowledge of this parameter for individual dosimeters allows to account for the variations of the conditions of calibration

  19. Need for improved standards in neutron personnel dosimetry

    International Nuclear Information System (INIS)

    Auxier, J.A.

    1976-01-01

    There is a continuing need for standards in neutron monitoring. A discussion of special problem areas and the benefits of intercomparisons is given. The RBE for leukemia induction in the survivors of the nuclear bombings of Hiroshima and Nagasaki is greater than ten for absorbed doses in the bone marrow of less than 100 rads; this may have an important impact on neutron standards preparation

  20. Neutron Activation Foil and Thermoluminescent Dosimeter Responses to a Lead Reflected Pulse of the CEA Valduc SILENE Critical Assembly

    International Nuclear Information System (INIS)

    Miller, Thomas Martin; Celik, Cihangir; Isbell, Kimberly McMahan; Lee, Yi-kang; Gagnier, Emmanuel; Authier, Nicolas; Piot, Jerome; Jacquet, Xavier; Rousseau, Guillaume; Reynolds, Kevin H.

    2016-01-01

    This benchmark experiment was conducted as a joint venture between the US Department of Energy (DOE) and the French Commissariat à l'Energie Atomique (CEA). Staff at the Oak Ridge National Laboratory (ORNL) in the US and the Centre de Valduc in France planned this experiment. The experiment was conducted on October 13, 2010 in the SILENE critical assembly facility at Valduc. Several other organizations contributed to this experiment and the subsequent evaluation, including CEA Saclay, Lawrence Livermore National Laboratory (LLNL), the Y-12 National Security Complex (NSC), Babcock International Group in the United Kingdom, and Los Alamos National Laboratory (LANL). The goal of this experiment was to measure neutron activation and thermoluminescent dosimeter (TLD) doses from a source similar to a fissile solution critical excursion. The resulting benchmark can be used for validation of computer codes and nuclear data libraries as required when performing analysis of criticality accident alarm systems (CAASs). A secondary goal of this experiment was to qualitatively test performance of two CAAS detectors similar to those currently and formerly in use in some US DOE facilities. The detectors tested were the CIDAS MkX and the Rocky Flats NCD-91. The CIDAS detects gammas with a Geiger-Muller tube, and the Rocky Flats detects neutrons via charged particles produced in a thin 6 LiF disc, depositing energy in a Si solid-state detector. These detectors were being evaluated to determine whether they would alarm, so they were not expected to generate benchmark quality data.

  1. Neutron Activation Foil and Thermoluminescent Dosimeter Responses to a Polyethylene Reflected Pulse of the CEA Valduc SILENE Critical Assembly

    Energy Technology Data Exchange (ETDEWEB)

    Miller, Thomas Martin [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Celik, Cihangir [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); McMahan, Kimberly L. [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Lee, Yi-kang [French Atomic Energy Commission (CEA), Saclay (France); Gagnier, Emmanuel [French Atomic Energy Commission (CEA), Centre de Saclay, Gif sur Yvette; Authier, Nicolas [French Atomic Energy Commission (CEA), Salives (France). Valduc Centre for Nuclear Studies; Piot, Jerome [French Atomic Energy Commission (CEA), Salives (France). Valduc Centre for Nuclear Studies; Jacquet, Xavier [French Atomic Energy Commission (CEA), Salives (France). Valduc Centre for Nuclear Studies; Rousseau, Guillaume [French Atomic Energy Commission (CEA), Salives (France). Valduc Centre for Nuclear Studies; Reynolds, Kevin H. [Y-12 National Security Complex, Oak Ridge, TN (United States)

    2016-09-01

    This benchmark experiment was conducted as a joint venture between the US Department of Energy (DOE) and the French Commissariat à l'Energie Atomique (CEA). Staff at the Oak Ridge National Laboratory (ORNL) in the US and the Centre de Valduc in France planned this experiment. The experiment was conducted on October 19, 2010 in the SILENE critical assembly facility at Valduc. Several other organizations contributed to this experiment and the subsequent evaluation, including CEA Saclay, Lawrence Livermore National Laboratory (LLNL), the Y-12 National Security Complex (NSC), Babcock International Group in the United Kingdom, and Los Alamos National Laboratory (LANL). The goal of this experiment was to measure neutron activation and thermoluminescent dosimeter (TLD) doses from a source similar to a fissile solution critical excursion. The resulting benchmark can be used for validation of computer codes and nuclear data libraries as required when performing analysis of criticality accident alarm systems (CAASs). A secondary goal of this experiment was to qualitatively test performance of two CAAS detectors similar to those currently and formerly in use in some US DOE facilities. The detectors tested were the CIDAS MkX and the Rocky Flats NCD-91. The CIDAS detects gammas with a Geiger-Muller tube and the Rocky Flats detects neutrons via charged particles produced in a thin 6LiF disc depositing energy in a Si solid state detector. These detectors were being evaluated to determine whether they would alarm, so they were not expected to generate benchmark quality data.

  2. Neutron Activation Foil and Thermoluminescent Dosimeter Responses to a Polyethylene Reflected Pulse of the CEA Valduc SILENE Critical Assembly

    International Nuclear Information System (INIS)

    Miller, Thomas Martin; Celik, Cihangir; McMahan, Kimberly L.; Lee, Yi-kang; Gagnier, Emmanuel; Authier, Nicolas

    2016-01-01

    This benchmark experiment was conducted as a joint venture between the US Department of Energy (DOE) and the French Commissariat à l'Energie Atomique (CEA). Staff at the Oak Ridge National Laboratory (ORNL) in the US and the Centre de Valduc in France planned this experiment. The experiment was conducted on October 19, 2010 in the SILENE critical assembly facility at Valduc. Several other organizations contributed to this experiment and the subsequent evaluation, including CEA Saclay, Lawrence Livermore National Laboratory (LLNL), the Y-12 National Security Complex (NSC), Babcock International Group in the United Kingdom, and Los Alamos National Laboratory (LANL). The goal of this experiment was to measure neutron activation and thermoluminescent dosimeter (TLD) doses from a source similar to a fissile solution critical excursion. The resulting benchmark can be used for validation of computer codes and nuclear data libraries as required when performing analysis of criticality accident alarm systems (CAASs). A secondary goal of this experiment was to qualitatively test performance of two CAAS detectors similar to those currently and formerly in use in some US DOE facilities. The detectors tested were the CIDAS MkX and the Rocky Flats NCD-91. The CIDAS detects gammas with a Geiger-Muller tube and the Rocky Flats detects neutrons via charged particles produced in a thin "6LiF disc depositing energy in a Si solid state detector. These detectors were being evaluated to determine whether they would alarm, so they were not expected to generate benchmark quality data.

  3. Magnetic field dosimeter development

    International Nuclear Information System (INIS)

    Lemon, D.K.; Skorpik, J.R.; Eick, J.L.

    1980-09-01

    In recent years there has been increased concern over potential health hazards related to exposure of personnel to magnetic fields. If exposure standards are to be established, then a means for measuring magnetic field dose must be available. To meet this need, the Department of Energy has funded development of prototype dosimeters at the Battelle Pacific Northwest Laboratory. This manual reviews the principle of operation of the dosimeter and also contains step-by-step instructions for its operation

  4. The development of an automatic scanning method for CR-39 neutron dosimeter

    International Nuclear Information System (INIS)

    Tawara, Hiroko; Miyajima, Mitsuhiro; Sasaki, Shin-ichi; Hozumi, Ken-ichi

    1989-01-01

    A method of measuring low level neutron dose has been developed with CR-39 track detectors using an automatic scanning system. It is composed of the optical microscope with a video camera, an image processor and a personal computer. The focus point of the microscope and the X-Y stage are controlled from the computer. The minimum detectable neutron dose is estimated at 4.6 mrem in the uniform field of neutron with equivalent energy spectrum to Am-Be source from the results of automatic measurements. (author)

  5. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford's mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue

  6. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    Energy Technology Data Exchange (ETDEWEB)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford`s mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue.

  7. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    Energy Technology Data Exchange (ETDEWEB)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford's mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue.

  8. Use of borated polyethylene to improve low energy response of a prompt gamma based neutron dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Priyada, P.; Ashwini, U.; Sarkar, P.K., E-mail: pradip.sarkar@manipal.edu

    2016-05-21

    The feasibility of using a combined sample of borated polyethylene and normal polyethylene to estimate neutron ambient dose equivalent from measured prompt gamma emissions is investigated theoretically to demonstrate improvements in low energy neutron dose response compared to only polyethylene. Monte Carlo simulations have been carried out using the FLUKA code to calculate the response of boron, hydrogen and carbon prompt gamma emissions to mono energetic neutrons. The weighted least square method is employed to arrive at the best linear combination of these responses that approximates the ICRP fluence to dose conversion coefficients well in the energy range of 10{sup −8} MeV to 14 MeV. The configuration of the combined system is optimized through FLUKA simulations. The proposed method is validated theoretically with five different workplace neutron spectra with satisfactory outcome. - Highlights: • An improved method is proposed for estimating H⁎(10) using prompt gamma emissions. • A combination of BHDPE and HDPE cylinders is used as a sample. • Linear combination of prompt gamma intensities approximates ICRP-DCC closely. • Feasibility of the method was tested theoretically using workplace neutron spectra.

  9. Personnel monitoring

    International Nuclear Information System (INIS)

    1965-01-01

    This film stresses the need for personnel monitoring in work areas where there is a hazard of exposure to radiation. It illustrates the use of personnel monitoring devices (specially the film dosimeter), the assessment of exposure to radiation and the detailed recording of the results on personnel filing cards

  10. Personnel monitoring

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1966-12-31

    This film stresses the need for personnel monitoring in work areas where there is a hazard of exposure to radiation. It illustrates the use of personnel monitoring devices (specially the film dosimeter), the assessment of exposure to radiation and the detailed recording of the results on personnel filing cards

  11. Dose-equivalent response CR-39 track detector for personnel neutron dosimetry

    International Nuclear Information System (INIS)

    Oda, K.; Ito, M.; Yoneda, H.; Miyake, H.; Yamamoto, J.; Tsuruta, T.

    1991-01-01

    A dose-equivalent response detector based on CR-39 has been designed to be applied for personnel neutron dosimetry. The intrinsic detection efficiency of bare CR-39 was first evaluated from irradiation experiments with monoenergetic neutrons and theoretical calculations. In the second step, the radiator effect was investigated for the purpose of sensitization to fast neutrons. A two-layer radiator consisting of deuterized dotriacontane (C 32 D 66 ) and polyethylene (CH 2 ) was designed. Finally, we made the CR-39 detector sensitive to thermal neutrons by doping with orthocarbone (B 10 H 12 C 2 ), and also estimated the contribution of albedo neutrons. It was found that the new detector - boron-doped CR-39 with the two-layer radiator - would have a flat response with an error of about 70% in a wide energy region, ranging from thermal to 15 MeV. (orig.)

  12. Personnel photon dosimeter on the base of TLD sup n sup a sup t LiF

    CERN Document Server

    Kaskanov, G Y

    2003-01-01

    A personnel photon dosemeter on the basis of thermoluminescence detectors (TLD) sup n sup a sup t LiF is described. Experimental responses of the dosemeters in the unit of individual equivalent dose H sub p (10) for energy of photons 59.5, 120, 662, and 1250 keV are presented. It is shown that the dosemeter allows one to measure H sub p (10) with admissible uncertainty in the energy range from 60 to 1250 keV.

  13. Neutron Activation Foil and Thermoluminescent Dosimeter Responses to a Lead Reflected Pulse of the CEA Valduc SILENE Critical Assembly

    Energy Technology Data Exchange (ETDEWEB)

    Miller, Thomas Martin [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Celik, Cihangir [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Isbell, Kimberly McMahan [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Lee, Yi-kang [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Gagnier, Emmanuel [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Authier, Nicolas [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Piot, Jerome [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Jacquet, Xavier [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Rousseau, Guillaume [Commissariat a l' Energie Atomique et aux Energies Alternatives (CEA-Saclay), Gif-sur-Yvette (France); Reynolds, Kevin H. [Y-12 National Security Complex, Oak Ridge, TN (United States)

    2016-09-01

    This benchmark experiment was conducted as a joint venture between the US Department of Energy (DOE) and the French Commissariat à l'Energie Atomique (CEA). Staff at the Oak Ridge National Laboratory (ORNL) in the US and the Centre de Valduc in France planned this experiment. The experiment was conducted on October 13, 2010 in the SILENE critical assembly facility at Valduc. Several other organizations contributed to this experiment and the subsequent evaluation, including CEA Saclay, Lawrence Livermore National Laboratory (LLNL), the Y-12 National Security Complex (NSC), Babcock International Group in the United Kingdom, and Los Alamos National Laboratory (LANL). The goal of this experiment was to measure neutron activation and thermoluminescent dosimeter (TLD) doses from a source similar to a fissile solution critical excursion. The resulting benchmark can be used for validation of computer codes and nuclear data libraries as required when performing analysis of criticality accident alarm systems (CAASs). A secondary goal of this experiment was to qualitatively test performance of two CAAS detectors similar to those currently and formerly in use in some US DOE facilities. The detectors tested were the CIDAS MkX and the Rocky Flats NCD-91. The CIDAS detects gammas with a Geiger-Muller tube, and the Rocky Flats detects neutrons via charged particles produced in a thin 6LiF disc, depositing energy in a Si solid-state detector. These detectors were being evaluated to determine whether they would alarm, so they were not expected to generate benchmark quality data.

  14. Development of integrated-type dosimeter responsive to high energy neutrons (2)

    Energy Technology Data Exchange (ETDEWEB)

    Sawamura, Teruko; Murai, Ikuo; Abe, Masashi; Uoyama, Kazuya; Das, Mala [Hokkaido Univ., Sapporo, Hokkaido (Japan); Tuda, Shuichi [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    2003-03-01

    The response of superheated drop detectors or bubble detectors (BDs) was measured for quasi-monoenergetic neutron beams in the 40-75 MeV range. The experiments were performed at the AVF cyclotron facility, TAKASAKI Ion Accelerator for Advanced Radiation Application (TIARA) of Japan Atomic Energy Research Institute (JAERI). The measured dose sensitivities showed to be lowered to about a half the nominal sensitivity. A lead-breeder introduced to extend response to the high energy region were investigated and compared with Monte Carlo calculations by MCNPX code. (author)

  15. Radiation dosimeter

    International Nuclear Information System (INIS)

    Lowe, D.

    1980-01-01

    A radiation dosimeter is described, comprising a thermoluminescent phosphor incorporated in matrix of polyethersulphone. The dosimeter is preferably a thin film formed by spreading a suspension of a powdered phosphor in a solution of polyethersulphone onto a flat surface. The solvent for the polyethersulphone is a mixture of a n-methyl-2-pyrrolidone and xylene in equal proportions. A thin, inert film of polyethersulphone can be cemented to one surface of the dosimeter so as to provide a skin dosimeter. (author)

  16. Plastic dosimeter

    International Nuclear Information System (INIS)

    Nagai, Shiro; Matsuda, Kohji.

    1988-01-01

    The report outlines major features and applications of plastic dosimeters. Some plastic dosimeters, including the CTA and PVC types, detect the response of the plastic material itself to radiations while others, such as pigment-added plastic dosimeters, contain additives as radiation detecting material. Most of these dosimeters make use of color centers produced in the dosimeter by radiations. The PMMA dosimeter is widely used in the field of radiation sterilization of food, feed and medical apparatus. The blue cellophane dosimeter is easy to handle if calibrated appropriately. The rad-color dosimeter serves to determine whether products have been irradiated appropriately. The CTA dosimeter has better damp proofing properties than the blue cellophane type. The pigment-added plastic dosimeter consists of a resin such as nylon, CTA or PVC that contains a dye. Some other plastic dosimeters are also described briefly. Though having many advantages, these plastic dosimeter have disadvantages as well. Some of their major disadvantages, including fading as well as large dependence on dose, temperature, humidity and anviroment, are discussed. (Nogami, K.)

  17. Kevlar® as a Potential Accident Radiation Dosimeter for First Responders, Law Enforcement and Military Personnel.

    Science.gov (United States)

    Romanyukha, Alexander; Trompier, François; Benevides, Luis A

    2016-08-01

    Today the armed forces and law enforcement personnel wear body armor, helmets, and flak jackets composed substantially of Kevlar® fiber to prevent bodily injury or death resulting from physical, ballistic, stab, and slash attacks. Therefore, there is a high probability that during a radiation accident or its aftermath, the Kevlar®-composed body armor will be irradiated. Preliminary study with samples of Kevlar® foundation fabric obtained from body armor used by the U.S. Marine Corps has shown that all samples evaluated demonstrated an EPR signal, and this signal increased with radiation dose. Based on these results, the authors predict that, with individual calibration, exposure at dose above 1 Gy can be reliably detected in Kevlar® samples obtained from body armor. As a result of these measurements, a post-event reconstruction of exposure dose can be obtained by taking various samples throughout the armor body and helmet worn by the same irradiated individual. The doses can be used to create a whole-body dose map that would be of vital importance in a case of a partial body or heterogeneous exposure.

  18. Personnel-dosimetry intercomparison studies at the Oak Ridge National Laboratory

    International Nuclear Information System (INIS)

    Swaja, R.E.; Sims, C.S.

    1982-01-01

    Since 1974, seven annual personnel dosimetry intercomparison studies have been conducted at the Oak Ridge National Laboratory using the Health Physics Reactor. These studies have produced more than 2000 measurements by 72 participating organizations of neutron and gamma dose equivalents between 0.1 to 15.0 mSv in six mixed radiation fields. The relative performance of three basic types of personnel neutron dosimeters (nuclear emulsion film, thermoluminescent, and track-etch) and two basic types of gamma dosimeters (film and thermoluminescent) was assessed based on experimental results obtained during the seven intercomparisons

  19. Applicability of thermoluminescent dosimeters in X-ray organ dose determination and in the dosimetry of systemic and boron neutron capture radiotherapy

    International Nuclear Information System (INIS)

    Aschan, C.

    1999-01-01

    The main detectors used for clinical dosimetry are ionisation chambers and semiconductors. Thermoluminescent (TL) dosimeters are also of interest because of their following advantages: (i) wide useful dose range, (ii) small physical size, (iii) no need for high voltage or cables, i.e. stand alone character, and (iv) tissue equivalence (LiF) for most radiation types. TL detectors can particularly be used for the absorbed dose measurements performed with the aim to investigate cases where dose prediction is difficult and not as part of a routine verification procedure. In this thesis, the applicability of TL detectors was studied in different clinical applications. Particularly, the major phenomena (e.g. energy dependence, sensitivity to high LET radiation, reproducibility) affecting on the precision and accuracy of TL detectors in the dose estimations were considered in this work. In organ dose determinations of diagnostic X-ray examinations, the TL detectors were found to be accurate within 5% (1 S.D.). For in viva studies using internal irradiation source, i.e. for systemic radiation therapy, a method for determining the absorbed doses to organs was introduced. The TL method developed was found to be able to estimate the absorbed doses to those critical organs near the body surface within 50%. In the mixed neutron-gamma field of boron neutron capture therapy (BNCT), TL detectors were used for gamma dose and neutron fluence measurements. They were found able to measure the neutron dose component with the accuracy of 16%, and therefore to be a useful addition to the activation foils in BNCT neutron dosimetry. The absorbed gamma doses can be measured with TL detectors within 20% in the mixed neutron-gamma field, which enables in viva measurements at BNCT beams with approximately the same accuracy. In this study, the uncertainties of TL dosimeters were found to be high but not essentially greater than those in other measurement techniques used for clinical dosimetry

  20. Pen dosimeters

    CERN Multimedia

    SC/RP Group

    2006-01-01

    The Radiation Protection Group has decided to withdraw all pen dosimeters from the main PS and SPS access points. This will be effective as of January 2006. The following changes will be implemented: All persons working in a limited-stay controlled radiation area must wear an operational dosimeter in addition to their personal DIS dosimeter. Any persons not equipped with this additional dosimeter must contact the SC/RP Group, which will make this type of dosimeter available for temporary loan. A notice giving the phone numbers of the SC/RP Group members to contact will be displayed at the former distribution points for the pen dosimeters. Thank you for your cooperation. The SC/RP Group

  1. Present status of fast neutron personnel dosimetry system based on CR-39 solid state nuclear track detectors

    International Nuclear Information System (INIS)

    Pal, Rupali; Sathian, Deepa; Jayalakshmi, V.; Bakshi, A.K.; Chougaonkar, M.P.; Mayya, Y.S.; Kumar, Valli; Babu, Rajesh; Kar, S.; Joshi, V.M.

    2011-08-01

    Neutron sources are of different types depending upon the method of production such as nuclear reactors, particle accelerators and laboratory sources. Neutron sources depending upon their energy, flux, size etc. are used for variety of applications in basic and applied sciences, neutron scattering experiments and in industry such as oil well - digging, coal mining and processing, ore processing etc. Personnel working in nuclear installations such as reactors, accelerators, spent fuel processing plants, nuclear fuel cycle operations and those working in various industries such as oil refining, oil well-digging, coal mining and processing, ore processing, etc. need to be monitored for neutron exposures, if any. Neutron monitoring is especially necessary in view of the fact that the radiation weighting factor for neutron is much higher than gamma rays and also it varies with energy. Radiological Physics and Advisory Division is involved in monitoring of personnel working in neutron fields. Around 2100 workers from 70 institutions (DAE and Non-DAE) are monitored on a quarterly basis. Neutron personnel monitoring, carried out in the country is based on Solid State Nuclear Track Detection (SSNTD) technique. In this technique, neutrons interact with hydrogen in CR-39 polymer to produce recoil protons. These protons create damages in the polymer, which are enlarged and appear as tracks when subjected to electrochemical etching (ECE). These tracks are counted in an optical system to evaluate the neutron dose. The neutron dosimetry system based on SSNTD has undergone a significant development, since it was started in 1990. The development includes upgradation of image analysis system for counting tracks, introduction of chemical etching (CE) at elevated temperatures for evaluation of dose equivalents above 10 mSv and use of carbon laser for cutting of CR-39 detectors. The entire dose evaluation process has been standardized, which includes calibration and performance tests

  2. Survey of international personnel radiation dosimetry programs

    International Nuclear Information System (INIS)

    Swaja, R.E.

    1985-04-01

    In September of 1983, a mail survey was conducted to determine the status of external personnel gamma and neutron radiation dosimetry programs at international agencies. A total of 130 agencies participated in this study including military, regulatory, university, hospital, laboratory, and utility facilities. Information concerning basic dosimeter types, calibration sources, calibration phantoms, corrections to dosimeter responses, evaluating agencies, dose equivalent reporting conventions, ranges of typical or expected dose equivalents, and degree of satisfaction with existing systems was obtained for the gamma and neutron personnel monitoring programs at responding agencies. Results of this survey indicate that to provide the best possible occupational radiation monitoring programs and to improve dosimetry accuracy in performance studies, facility dosimetrists, regulatory and standards agencies, and research laboratories must act within their areas of responsibility to become familiar with their radiation monitoring systems, establish common reporting guidelines and performance standards, and provide opportunities for dosimetry testing and evaluation. 14 references, 10 tables

  3. Cell-phone interference with pocket dosimeters

    International Nuclear Information System (INIS)

    Djajaputra, David; Nehru, Ramasamy; Bruch, Philip M; Ayyangar, Komanduri M; Raman, Natarajan V; Enke, Charles A

    2005-01-01

    Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag. (note)

  4. Cell-phone interference with pocket dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Djajaputra, David; Nehru, Ramasamy; Bruch, Philip M; Ayyangar, Komanduri M; Raman, Natarajan V; Enke, Charles A [Department of Radiation Oncology, University of Nebraska Medical Center, 987521 Nebraska Medical Center, Omaha, NE 68198-7521 (United States)

    2005-05-07

    Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag. (note)

  5. The role of phantom parameters on the response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter

    International Nuclear Information System (INIS)

    Sohrabi, M.; Katouzi, M.

    1992-01-01

    The response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter (NANPD) which can also be used for other albedo dosemeter types was determined on 18 different phantom configurations. The effects of type, geometry, material, thickness, dosemeter-to-phantom angle in particular with the presence of legs were investigated using a Pu-Be neutron source. It was concluded that the slab phantoms (single or double) and circular and elliptical cylinder phantoms seemed to provide a better response, whereas the ICRU sphere geometry does not seem to be appropriate for the calibration of albedo dosemeters. It is interesting to note that the presence of legs maintains the constancy of the response in a situation when a radiation worker bends down during work. (author)

  6. Retrospective assessment of personnel neutron dosimetry for workers at the Hanford Site

    International Nuclear Information System (INIS)

    Fix, J.J.; Wilson, R.H.; Baumgartner, W.B.

    1996-09-01

    This report was prepared to examine the specific issue of the potential for unrecorded neutron dose for Hanford workers, particularly in comparison with the recorded whole body (neutron plus photon) dose. During the past several years, historical personnel dosimetry practices at Hanford have been documented in several technical reports. This documentation provides a detailed history of the technology, radiation fields, and administrative practices used to measure and record dose for Hanford workers. Importantly, documentation has been prepared by personnel whose collective experience spans nearly the entire history of Hanford operations beginning in the mid-1940s. Evaluations of selected Hanford radiation dose records have been conducted along with statistical profiles of the recorded dose data. The history of Hanford personnel dosimetry is complex, spanning substantial evolution in radiation protection technology, concepts, and standards. Epidemiologic assessments of Hanford worker mortality and radiation dose data were initiated in the early 1960s. In recent years, Hanford data have been included in combined analyses of worker cohorts from several Department of Energy (DOE) sites and from several countries through the International Agency for Research on Cancer (IARC). Hanford data have also been included in the DOE Comprehensive Epidemiologic Data Resource (CEDR). In the analysis of Hanford, and other site data, the question of comparability of recorded dose through time and across the respective sites has arisen. DOE formed a dosimetry working group composed of dosimetrists and epidemiologists to evaluate data and documentation requirements of CEDR. This working group included in its recommendations the high priority for documentation of site-specific radiation dosimetry practices used to measure and record worker dose by the respective DOE sites

  7. Personnel Dosimeters with Radiation Elements; Dosimetres Individuels Comportant des Elements Emetteurs; 0418 041d 0414 0418 0412 0418 0414 ; Dosimetros Individuales Provistos de Elementos Emisores

    Energy Technology Data Exchange (ETDEWEB)

    Hosemann, R.; Warrikhoff, H. F.H. [Fritz-Haber-Institut der Max-Planck-Gesellschaft, Berlin-Dahlem, Federal Republic of Germany (Germany)

    1965-06-15

    The radiation element is a vacuum diode, the electrodes of which emit a different charge of electrons per unit of irradiation. Since the electrons ejected leave one electrode with a high energy, the other electrode will be charged to proportionally high voltages without any auxiliary power. Accordingly these direct reading systems are called radiation elements. When measuring X- and y-rays, one electrode consists of lead, the other of carbon. The wavelength independence with regard to the r-unit was obtained by the electron-filtering method. The sensitivity is approximately 0.5 V/r {+-} 8% within the range from 80 keV to 1.25 MeV. (Full scale 200-2000 r either linear or quasi logarithm. ) When measuring slow and fast neutrons, the induced {beta}-activity of the electrodes charge the built-in electrometer. It is possible to measure both gamma radiation and neutrons in rem units with one instrument. The neutron sensitivity of a gamma element can be made definitely zero, but the gamma sensitivity of a neutron element presents some difficulties. The outstanding advantages of these instruments are: (1) Self-powered systems, always ready to function, unlimited storing time, no batteries or charging units; (2) Extremely low fading because of the use of vacuum and quartz for isolation; (3) Quantitative measurement of the highest attainable dose rates because of no recombination effects (vacuum) ; and (4) Either direct or indirect systems. The direct reading instrument contains the radiation element and the electrometer in the same vacuum vessel. There are no electrical contacts outside the instrument. Cylindrical and spherical designs of the element are preferable to other geometries. The outer dimensions of the instrument will be the same as those of the commonly used pen-type dosimeters. Each system contains a discharge switch and a control unit. (author) [French] L'element emetteur est constitue par une diode a vide dont les electrodes emettent des flux d

  8. Neutron measurements at nuclear power reactors [55

    CERN Document Server

    Scherpelz, R I

    2002-01-01

    Staff from the Pacific Northwest National Laboratory (operated by Battelle Memorial Institute), have performed neutron measurements at a number of commercial nuclear power plants in the United States. Neutron radiation fields at light water reactor (LWR) power plants are typically characterized by low-energy distributions due to the presence of large amounts of scattering material such as water and concrete. These low-energy distributions make it difficult to accurately monitor personnel exposures, since most survey meters and dosimeters are calibrated to higher-energy fields such as those produced by bare or D sub 2 O-moderated sup 2 sup 5 sup 2 Cf sources. Commercial plants typically use thermoluminescent dosimeters in an albedo configuration for personnel dosimetry and survey meters based on a thermal-neutron detector inside a cylindrical or spherical moderator for dose rate assessment, so their methods of routine monitoring are highly dependent on the energy of the neutron fields. Battelle has participate...

  9. Interim status report of the TMI personnel-dosimetry project

    International Nuclear Information System (INIS)

    Rich, B.L.; Alvarez, J.L.; Adams, S.R.

    1981-06-01

    The current 2-chip TLD personnel dosimeter in use at Three Mile Island (TMI) has been shown inadequate for the anticipated high beta/gamma fields during TMI recovery operations in some areas. This project surveyed the available dosimeter systems, set up an Idaho National Engineering Laboratory (INEL) prototype system, and compared this system with those commercial systems that could be made immediately available for comparison. Of the systems tested, the new INEL personnel dosimeter was found to produce the most accurate results for use in recovery operations at TMI-2. The other multiple-chip or multiple-filter systems were found less desirable at present. The most prominent deficiencies in the INEL dosimeter stem from the fact that it lacks a completely automated reader and its x-ray and thermal neutron responses require additional development. A automated prototype reader system may be in operation by the end of CY-1981. Three alternatives for operational dosimetry are discussed. A combination of a modified version of the presently used Harshaw 2-chip dosimeter and the INEL dosimeter is recommended

  10. Setting-Up of a Neutron Dosimeter Using Ionographic Tracks in an Insulating Solid; Realisation d'un Dosimetre a Neutrons Utilisant les Traces Ionographiques dans un Solide Isolant

    Energy Technology Data Exchange (ETDEWEB)

    Carfi, N.; Dugnani Lonati, R.; Mangialajo, M. [Centro Informazioni Studi Esperienze, Segrate, Milan (Italy)

    1971-06-15

    The use of dielectrics such as mica, glasses and plastic materials for charged particle track recording has been developed in the past years; track recording of fission fragments produced in fissile materials under neutron irradiation is particularly interesting. The method has not as yet been applied to neutron dosimetry owing to the difficulties of track observation and counting using optical microscopes. A method has recently been proposed which results in markedly reduced counting times, electronic counting of the electric discharges between two electrodes through the dielectric which was perforated by the fission fragments. The reasons for adopting this new technique to set-up a neutron dosimeter are: (a) economic advantages, since a rather simple electronic instrumentation will give a reading of the dosimeter, preparation of which can be carried out immediately and at a modest cost; (b) fast reading, since the response can be obtained in less than two hours from irradiation; (c) dosimetric advantages owing to the lack of fading and insensitivity to {beta} or {gamma} radiations. The present work describes the method adopted and the equipment used to set up the dosimeter. Results are reported and briefly discussed. (author) [French] Au cours des dernieres annees on a assiste a un developpement remarquable de l' utilisation d' isolants tels que le mica, les verres et les materiaux plastiques pour l' enregistrement des traces de particules chargees; l' enregistrement des traces de fragments de fission produits dans les materiaux fissiles sous irradiation neutronique est particulierement interessant. Cette methode n'a pas, jusqu a present, ete appliquee a la dosimetrie des neutrons a cause des difficultes que presentent l' observation et le comptage des traces au microscope optique. On a propose recemment une methode qui permet de reduire considerablement la duree de l' operation en comptant electroniquement les decharges electriques qui se produisent entre deux

  11. High-energy neutron dosimetry at the Clinton P. Anderson Meson Physics Facility

    International Nuclear Information System (INIS)

    Mallett, M.W.; Vasilik, D.G.; Littlejohn, G.J.; Cortez, J.R.

    1990-01-01

    Neutron energy spectrum measurements performed at the Clinton P. Anderson Meson Physics Facility indicated potential areas for high energy neutron exposure to personnel. The low sensitivity of the Los Alamos thermoluminescent dosimeter (TLD) to high energy neutrons warranted issuing a NTA dosimeter in addition to the TLD badge to employees entering these areas. The dosimeter consists of a plastic holder surrounding NTA film that has been desiccated and sealed in a dry nitrogen environment. A study of the fading of latent images in NTA film demonstrated the success of this packaging method to control the phenomenon. The Los Alamos NTA dosimeter is characterized and the fading study discussed. 10 refs., 4 figs., 2 tabs

  12. Chemical dosimeter

    International Nuclear Information System (INIS)

    Baker, W.B.; Clark, D.G.

    1979-01-01

    The dosimeter may be carried by individuals e.g. at the belt and serves to monitor for vinyl-chloride vapors in industrial plants and for toxic radon gas and toxic radon gas products in mines. It contains a pump, sucking an air flow through an orifice and a filter, as well as a sensor circuit for detecting low air flow rates and a battery testing circuit. (DG) 891 HP/DG 892 MKO [de

  13. Evaluation of different polymers for fast neutron personnel dosimetry using electrochemical etching

    International Nuclear Information System (INIS)

    Gammage, R.B.; Cotter, S.J.

    1977-01-01

    There is considerable optimism for the enhancement by electrochemical etching of fast neutron-induced recoil tracks in polycarbonate for the purpose of personnel dosimetry. The threshold energy, however, is rather high. A desirable improvement would be to lower this energy below 1 MeV. With this objective in mind, we have commenced an investigation of cellulose acetate, triacetate, and acetobutyrate in addition to polycarbonate. These cellulose derivatives are chemically more reactive and physically weaker than polycarbonate. It might, therefore, be possible to initiate the electrochemical amplification at the sites of shorter recoil atom damage tracks than is possible with polycarbonate. Some characteristics important for electrochemically etching in aqueous electrolytes are listed. Chemical etching is combined with treeing, an electrical breakdown process that starts when the dielectric strength is exceeded. These mechanical and electrical properties pertain to the dry plastics. The absorption of water molecules and electrolyte ions will cause these values to be reduced. Results and conclusions of the study are presented

  14. Composite material dosimeters

    Science.gov (United States)

    Miller, Steven D.

    1996-01-01

    The present invention is a composite material containing a mix of dosimeter material powder and a polymer powder wherein the polymer is transparent to the photon emission of the dosimeter material powder. By mixing dosimeter material powder with polymer powder, less dosimeter material is needed compared to a monolithic dosimeter material chip. Interrogation is done with excitation by visible light.

  15. Some aspects on neutron dosimetry

    International Nuclear Information System (INIS)

    Henaish, B.A.; Youssef, S.K.

    1988-01-01

    The American National Council on Radiation Protection and measurements (1) has recently issued a statement regarding dose limitation system for neutrons. The changes proposed in that statement presented substantial problems regarding the personnel exposure to neutrons and had pointed out the need to reassess an adequate current neutron dosimetry practice. Generally, the same types of dosimeters i.e. Nuclear Track (NTA films) and TLD-Albedo, have been used at major nuclear facilities over the past 15 years. here recently, other dosimetry methods such as track etch with polycarbonates such as CR-39 have been developed. However these should be recognized as local systems aiming to the development of better and more applicable dosimeters. 4 tab

  16. Nuclear radiation monitoring instruments for personnel in nuclear disaster for defence needs

    International Nuclear Information System (INIS)

    Bhatnagar, P.K.; Vaijapurkar, S.G.; Yadav, Ashok

    2005-01-01

    Ever since the tragedy of nuclear device exploding over Japan by USA in 1945 awareness exists amongst the armed forces personnel all over the world that a requirement of implementing radiological protection is imminent. Towards this adoption of radiological safety programme is a criterion. In a nuclear war disaster scenario, one encounters initial nuclear radiation (gamma and neutron radiations), gamma radiations from fallout, heat and blast. At certain distances Tanks/ armoured vehicles will survive and 4 R/s radiation level sensing to actuate relays for closing the ports of vehicles is essential, leading to reduction in inhalation, ingestion of fallout radioactivity and reduction in radiation dose received by occupants of the vehicle. Towards this sturdy radiation monitors to indicate gamma dose rate of the order of 1000 R/h, gamma and neutron dosimeters of the order of 1000 cGy with reading instruments are to be developed. These must work in harsh environment and sustain JSS 55555 conditions of army. Defence Laboratory, Jodhpur over past one decade has been involved in developing personnel, area and field monitoring instruments like dosimeters, survey meters, which are useful, acceptable to army personnel, armoured and personnel carrier vehicles, field structures/shelters. Technology transfer after satisfaction of armed forces, product ionisation and supply, maintenance, training has been the endeavor of the DRDO. Herein it is proposed to highlight the techno electronics nuclear radiation monitoring sensors and associated electronics systems developed first time in the country and productionised in bulk for Services for implementing personnel protection. The sensors developed and described are - Radiophotoluminescent Glass (RPLG) for gamma radiation dosimetry , neutron sensitive PIN diode for fast neutron dosimetry, gamma radiation sensitive PIN diode, superheated liquid neutron and gamma sensors. The dosimeter, dose rate meter and field/area instruments are

  17. Dose measurement during defectoscopic work using electronic personal dosimeters

    International Nuclear Information System (INIS)

    Smoldasova, J.

    2008-01-01

    Personal monitoring of the external radiation of radiation, personnel exposed to sources of ionizing radiation at a workplace is an important task of the radiological protection. Information based on the measured quantities characterizing the level of the exposure of radiation personnel enable to assess the optimum radiological protection at the relevant workplace and ascertain any deviation from the normal operation in time. Different types of personal dosimeters are used to monitor the external radiation of radiation personnel. Basically, there are two types of dosimeters, passive and active (electronic). Passive dosimeters provide information on the dose of exposure after its evaluation, while electronic dosimeters provide this information instantly. The goal of the work is to compare data acquired during different working activities using the DMC 2000 XB electronic dosimeters and the passive film dosimeters currently used at the defectoscopic workplace. (authors)

  18. Optical dosimeter

    International Nuclear Information System (INIS)

    Drukaroff, I.; Fishman, R.

    1984-01-01

    A reflecting optical dosimeter is a thin block of optical material having an input light pipe at one corner and an output light pipe at another corner, arranged so that the light path includes several reflections off the edges of the block to thereby greatly extend its length. In a preferred embodiment, one corner of the block is formed at an angle so that after the light is reflected several times between two opposite edges, it is then reflected several more times between the other two edges

  19. Performance comparisons of selected personnel-dosimetry systems in use at Department of Energy facilities

    International Nuclear Information System (INIS)

    Roberson, P.L; Holbrook, K.L.; Yoder, R.C.; Fox, R.A.; Hadley, R.T.; Hogan, B.T.; Hooker, C.D.

    1983-10-01

    Dosimeter performance data were collected to help develop a uniform approach to the calibration and use of personnel dosimetry systems for Department of Energy (DOE) laboratories. Eleven DOE laboratories participated in six months of testing using the American National Draft Standard, Criteria for Testing Personnel Dosimetry Performance, ANSI N13.11, and additional testing categories. The tests described in ANSI N13.11 used a pass/fail system to determine compliance with the draft standard. Recalculation to PNL irradiations showed that the 137 Cs, 90 Sr/ 90 Y, and 252 Cf categories can be recalibrated to have acceptable performance for nearly all participant systems. Deficient dosimeter design or handling techniques caused poor performance in the x-ray category for nearly half of the participants. Too little filtration for the deep-dose element caused poor performance in the beta/photon mixture category for one participant. Two participants had excessively high standard deviations in the neutron category due to dosimeter design or handling deficiencies. The participating dosimetry systems were separated into three categories on their dose evaluation procedure for low-energy photons. These were film dosimeters, fixed-calibration thermoluminescent (TL) dosimeters, and variable-calibration TL dosimeters. The performance of the variable-calibration design was best while the film dosimeters performed considerably worse than either TL dosimeter design. Beta energy dependence studies confirmed a strong correlation between sensitive element thickness, shallow element filtration and low-energy beta response. Studies of neutron calibration conditions for each participant suggested a relationship between response and calibration facility design

  20. Portable dosimeter

    International Nuclear Information System (INIS)

    Buffa, A.; Caley, R.; Pfaff, K.

    1986-01-01

    A simple but very accurate portable dosimeter is described for indicating the intensity of ionizing radiation, comprising, as a unit: (a) a radiation-detection chamber having a pair of parallel, facing, electrically-conducting, radiation-permeable electrodes spaced from each other to define a volume for a gas which is ionized by the radiation when exposed thereto; (b) electric potential supply means connected across the electrodes for attracting the gas ions to the electrodes and transferring their charge to the electrodes; (c) detection circuit means connected across the electrodes and having at least one of high-frequency electromagnetic- and radiation-sensitive components for detecting the charge on the electrodes and indicating therefrom a representation of the intensity of the radiation; (d) radiation shield means surrounding the radiation-sensitive components of the detection circuit means for shielding the latter from the ionizing radiation; (e) electric shield means surrounding the sensitive components of the detection circuit means for shielding the latter from electromagnetic interference including any caused by the ionizing radiation; and (f) ion shield means potting the ion-sensitive components for shielding them from radiation-caused ambient ionization; whereby the entire dosimeter may be assembled as the unit and portably transported into various radiation sources

  1. The effects of pre-etching time on the characteristic responses of electrochemically etched CR-39 neutron dosimeters

    International Nuclear Information System (INIS)

    Sohrabi, M.; Khoshnoodi, M.

    1986-01-01

    The effects of pre-etching time (PET) or duration of etching of fast-neutron-induced-recoil tracks in CR-39 in 6N KOH at 60 0 C on electrochemical etching neutron characteristic responses; i.e. sensitivity and mean recoil track diameter (MRTD) versus KOH normality up to 18N are investigated in this paper. Six sets of responses for PETs of 0, 1, 2, 3, 4, and 5 hours were obtained by using our new multi-chamber ECE (MCECE) system which reduced total operation time to about 6% of the time usually required when single-chamber ECE systems are used. The sensitivity response for zero PET showed a broad plateau and a high sensitivity low-LET peak around 16N. By increasing PET, another peak was also developed around 5N leading to 'double-humped' responses with two maximums around 5N and 16N, and a minimum around 11N. On the other hand, the MRTD responses for all PETs studied showed the same general trend with maximums around 11N. In this paper, shape of tracks under different conditions are also investigated, new optimum conditions such as KOH concentrations of 5, 11, and 15N at 25 0 C, with or without pre-etching, are recommended for tracks of lower-LET recoils including possibly protons, and alpha particle tracks over a broad energy range, and the efficiency of the MCECE system is also demonstrated. (author)

  2. Neutron response study

    International Nuclear Information System (INIS)

    Endres, G.W.R.; Fix, J.J.; Thorson, M.R.; Nichols, L.L.

    1981-01-01

    Neutron response of the albedo type dosimeter is strongly dependent on the energy of the incident neutrons as well as the moderating material on the backside of the dosimeter. This study characterizes the response of the Hanford dosimeter for a variety of neutron energies for both a water and Rando phantom (a simulated human body consisting of an actual human skeleton with plastic for body muscles and certain organs). The Hanford dosimeter response to neutrons of different energies is typical of albedo type dosimeters. An approximate two orders of magnitude difference in response is observed between neutron energies of 100 keV and 10 MeV. Methods were described to compensate for the difference in dosimeter response between a laboratory neutron spectrum and the different spectra encountered at various facilities in the field. Generally, substantial field support is necessary for accurate neutron dosimetry

  3. Fast fluence measurement for JOYO irradiation field using niobium dosimeter

    International Nuclear Information System (INIS)

    Ito, Chikara

    2004-03-01

    Neutron fluence and spectrum are key parameters in various irradiation tests and material surveillance tests so they need to be evaluated accurately. The reactor dosimetry test has been conducted by the multiple foil activation method, and a niobium dosimeter has been developed for measurement of fast neutron fluence in the experimental fast reactor JOYO. The inelastic scattering reaction of 93 Nb has a low threshold energy, about 30 keV, and the energy distribution of reaction cross section is similar to the displacement cross section for iron. Therefore, a niobium dosimeter is suitable for evaluation of the fast neutron fluence and the displacement per atom for iron. Moreover, a niobium dosimeter is suited to measure neutron fluence in long-term irradiation test because 93 Nb, which is produced by the reaction, has a long half-life (16.4 years). This study established a high precision measurement technique using the niobium reaction rate. The effect of self-absorption was decreased by the solution and evaporation to dryness of niobium dosimeter. The dosimeter weight was precisely measured using the inductively coupled plasma mass spectrometer. This technique was applied to JOYO dosimetry. The fast neutron fluences (E > 0.1 MeV) found by measuring the reaction rate in the niobium dosimeter were compared with the values evaluated using the multiple foil activation method. The ratio of measured fast neutron fluences by means of niobium dosimeter and multiple foil activation method range from 0.97 to 1.03 and agree within the experimental uncertainty. The measurement errors of fast neutron fluence by niobium dosimeter range from 4.5% (fuel region) to 10.1% (in-vessel storage rack). As a result of this study, the high precision measurement of fast neutron fluence by niobium dosimeters was confirmed. The accuracy of fast reactor dosimetry will be improved by application of niobium dosimeters to the irradiation tests in the JOYO MK-III core. (author)

  4. Neutron doses to personnel from a 24 MeV betatron

    International Nuclear Information System (INIS)

    Beckham, W.A; Entwistle, R.F.

    1987-01-01

    Neutrons are produced by bombardment of most materials by high-energy photons. Because the x-ray shielding around high-energy x-ray generators may not have been designed with neutrons in mind there may be unexpected contributions to the radiation doses of staff working in the immediate vicinity. Neutron fluxes in the working area close to an Allis-Chalmers 24 MeV betatron have been measured using a lithium-6-loaded scintillator and the dose rates calculated. Hazard of staff has been found to be low; typical dose-equivalent rates in occupied areas range from 0.0042 to 0.012 mrem/hour. The flux of fast neutrons in the treatment room was found to be essentially zero. Measurements of neutron flux may be routinely performed using the scintillation detector (NE 912) described, and could usefully form part of the acceptance protocol for any new accelerator

  5. Photon and neutron doses of the personnel using moisture and density measurement devices

    Energy Technology Data Exchange (ETDEWEB)

    Carinou, E.; Papadomarkaki, E.; Tritakis, P.; Hourdakis, C.I.; Kamenopoulou, V. [Greek Atomic Energy Commission, Agia Paraskevi, Attiki, 60092 (Greece)

    2006-07-01

    The objective of this study is to present the evolution of the photon doses received by the workers who use mobile devices for measuring the moisture and the density in various materials and to estimate the neutron doses. The workers employed in more than 30 construction companies in Greece were 76 in 2004. The devices used for that purpose incorporate a {sup 137}Cs source for density measurements and an {sup 241}Am-Be source for moisture measurements of soil, asphalt or concrete. Photon and neutron measurements were performed occasionally during the on site inspections. The results of the measurements showed that the photon and neutron dose rates were not negligible. The workers were monitored for photon radiation using film badges (Kodak Type 2, Holder NRPB type) till the year 2000 and then TLD badges issued by the Greek Atomic Energy Commission (GAEC), on a monthly basis. Since the neutron dose rates measured by a rem-meter were not so high, no neutron dosemeters were issued for them. Their personal dose equivalent data for photons are kept in the National Dose Registry Information System (N.D.R.I.S.) in G.A.E.C. and were used for statistical analysis for the period from 1997 till 2004. As far as the neutrons are concerned, a Monte Carlo code was used to simulate the measuring devices and the working positions in order to calculate the neutron individual doses. (authors)

  6. Photon and neutron doses of the personnel using moisture and density measurement devices

    International Nuclear Information System (INIS)

    Carinou, E.; Papadomarkaki, E.; Tritakis, P.; Hourdakis, C.I.; Kamenopoulou, V.

    2006-01-01

    The objective of this study is to present the evolution of the photon doses received by the workers who use mobile devices for measuring the moisture and the density in various materials and to estimate the neutron doses. The workers employed in more than 30 construction companies in Greece were 76 in 2004. The devices used for that purpose incorporate a 137 Cs source for density measurements and an 241 Am-Be source for moisture measurements of soil, asphalt or concrete. Photon and neutron measurements were performed occasionally during the on site inspections. The results of the measurements showed that the photon and neutron dose rates were not negligible. The workers were monitored for photon radiation using film badges (Kodak Type 2, Holder NRPB type) till the year 2000 and then TLD badges issued by the Greek Atomic Energy Commission (GAEC), on a monthly basis. Since the neutron dose rates measured by a rem-meter were not so high, no neutron dosemeters were issued for them. Their personal dose equivalent data for photons are kept in the National Dose Registry Information System (N.D.R.I.S.) in G.A.E.C. and were used for statistical analysis for the period from 1997 till 2004. As far as the neutrons are concerned, a Monte Carlo code was used to simulate the measuring devices and the working positions in order to calculate the neutron individual doses. (authors)

  7. Eleventh ORNL personnel dosimetry intercomparison study, May 22-23, 1985

    International Nuclear Information System (INIS)

    Swaja, R.E.; Oyan, R.; Sims, C.S.

    1986-07-01

    The Eleventh Personnel Dosimetry Intercomparison Study was conducted at the Oak Ridge National Laboratory (ORNL) during May 22-23, 1985. Dosimeter badges from 44 participating organizations were mounted on Lucite block phantoms and exposed to four mixed-radiation fields with neutron dose equivalents around 5 mSv and gamma dose equivalents between 0.1 and 0.7 mSv. Results of this study indicated that no participants had difficulty obtaining measurable indication of neutron exposure at the provided dose equivalent levels, and very few had difficulty obtaining indication of gamma exposure at dose equivalents as low as 0.10 mSv. Average neutron results for all dosimeter types were within 20% of reference values with no obvious spectrum dependence. Different dosimeter types (albedo, direct interaction TLD, film, recoil track, and combination albedo-track) with 10 or more reported measurements provided average results within 35% of reference values for all spectra. With regard to precision, about 80% of the reported neutron results had single standard deviations within 10% at the means which indicates that precision is not a problem relative to accuracy for most participants. Average gamma results were greater than reference values by factors of 1.07 to 1.52 for the four exposures with TLD systems being more accurate than film. About 80% of all neutron results and 67% of all gamma results met regulatory standards for measurement accuracy and approximately 70% of all neutron data satisfied national dosimetry accreditation criteria for accuracy plus precision. In general, neutron dosimeter performance observed in this intercomparison was much improved compared to that observed in the prior studies while gamma dosimeter performance was about the same

  8. LLL development of a combined etch track: albedo dosimeter

    International Nuclear Information System (INIS)

    Griffith, R.V.; Fisher, J.C.; Harder, C.A.

    1977-01-01

    The addition of polycarbonate sheet to albedo detectors for electrochemical etching provides a simple, inexpensive way to reduce the spectral sensitivity of the personnel dosimeter without losing the albedo features of sensitivity and ease of automation. The ECEP technique also provides the dosimetrist with the potential for identifying conditions of body orientation that might otherwise lead to significant error in dosimeter evaluation

  9. Evaluation of LLNL's Nuclear Accident Dosimeters at the CALIBAN Reactor September 2010

    International Nuclear Information System (INIS)

    Hickman, D.P.; Wysong, A.R.; Heinrichs, D.P.; Wong, C.T.; Merritt, M.J.; Topper, J.D.; Gressmann, F.A.; Madden, D.J.

    2011-01-01

    The Lawrence Livermore National Laboratory uses neutron activation elements in a Panasonic TLD holder as a personnel nuclear accident dosimeter (PNAD). The LLNL PNAD has periodically been tested using a Cf-252 neutron source, however until 2009, it was more than 25 years since the PNAD has been tested against a source of neutrons that arise from a reactor generated neutron spectrum that simulates a criticality. In October 2009, LLNL participated in an intercomparison of nuclear accident dosimeters at the CEA Valduc Silene reactor (Hickman, et.al. 2010). In September 2010, LLNL participated in a second intercomparison of nuclear accident dosimeters at CEA Valduc. The reactor generated neutron irradiations for the 2010 exercise were performed at the Caliban reactor. The Caliban results are described in this report. The procedure for measuring the nuclear accident dosimeters in the event of an accident has a solid foundation based on many experimental results and comparisons. The entire process, from receiving the activated NADs to collecting and storing them after counting was executed successfully in a field based operation. Under normal conditions at LLNL, detectors are ready and available 24/7 to perform the necessary measurement of nuclear accident components. Likewise LLNL maintains processing laboratories that are separated from the areas where measurements occur, but contained within the same facility for easy movement from processing area to measurement area. In the event of a loss of LLNL permanent facilities, the Caliban and previous Silene exercises have demonstrated that LLNL can establish field operations that will very good nuclear accident dosimetry results. There are still several aspects of LLNL's nuclear accident dosimetry program that have not been tested or confirmed. For instance, LLNL's method for using of biological samples (blood and hair) has not been verified since the method was first developed in the 1980's. Because LLNL and the other DOE

  10. Comparison of alanine dosimeters using silicone as their binder to a commercial, polystyrene-bound, alanine dosimeter

    International Nuclear Information System (INIS)

    Galindo, S.; Urena-Nunez, F.

    1997-01-01

    The feasibility of practical boron-containing alanine ESR dosimeters for gamma-neutron mixed field irradiation dosimeters depends in part on whether the γ response characteristics of these silicone-bound dosimeters are comparable to those of a commercially available dosimeter that has been used by the International Atomic Energy Agency (International Dose Assurance Service) as a transfer reference dosimeter. This work presents the results of the comparison of 3 batches of silicone-bound alanine dosimeters. The first batch consists of a mixture of alanine and boric acid; the second, alanine and borax; and the last contains only alanine. Results indicate that γ response characteristics of the silicone-bound samples are comparable to those of the commercial, polystyrene-bound, alanine dosimeter and that silicone has a strong potential as a binding substance for alanine ESR dosimetry. (Author)

  11. Operational aspects of the direct ion storage dosimeter system: 18 months of experience at CERN

    International Nuclear Information System (INIS)

    Carbonez, P.; Kotamaki, E.; Otto, Th.

    2006-01-01

    CERN, the European Organization for Nuclear Research, operates a dosimetry service for external exposure. The service monitors approximately 5000 Staff, scientific visitors and contractors personnel working on the organizations sites with personal dosimeters for personal dose equivalent (Hp(10), Hp(0.07)) from gamma, beta and neutron radiation. The dosimetry service is approved by the Swiss Federal Health Office, the competent authority for radiation protection. In 2004, the R.A.D.O.S. D.I.S. -1 dosemeter has been introduced to CERN as a gamma-beta dosemeter. Technical characteristics of this dosimeter, based on the direct ion storage technology, are high sensitivity, excellent linearity of the personal dose response with respect to radiation energy or dose, and long-term physical storage of personal dose-related information without the risk of fading. One important technical feature of the dosemeter is its 'instant reading' capability: the user himself can evaluate the received personal dose nondestructively on specific reader stations. This information is digitized, centralized by the CERN-wide network and stored in a database. The consequence of the 'instant reading' capability is a break with in the traditional organisation of a dosimetry service. The personal dosimeters are no longer exchanged periodically for evaluation, but a monthly value of personal dose is calculated from the readings initiated by the user. After a wearing period of one year, users are invited to exchange the dosimeter against a new, recently calibrated unit. The introduction of the D.I.S.-1 dosimeter has profoundly changed the type of work in CERN dosimetry service. Technical and laboratory work (development of film dosimeters, densitometric evaluation) have made place for computer-based procedures and database management. (authors)

  12. Radiation dosimeter utilizing the thermoluminescence of lithium fluoride.

    Science.gov (United States)

    CAMERON, J R; DANIELS, F; JOHNSON, N; KENNEY, G

    1961-08-04

    A dosimeter, with little wavelength dependence and large useful energy range for electromagnetic radiation, which is simple to use and read, has been developed. It appears to have applications in personnel monitoring as well as radiation research.

  13. Performance testing of extremity dosimeters, Study 2

    International Nuclear Information System (INIS)

    Harty, R.; Reece, W.D.; Hooker, C.D.

    1990-04-01

    The Health Physics Society Standards Committee (HPSSC) Working Group on Performance Testing of Extremity Dosimeters has issued a draft of a proposed standard for extremity dosimeters. The draft standard proposes methods to be used for testing dosimetry systems that determine occupational radiation dose to the extremities and the performance criterion used to determine compliance with the standard. Pacific Northwest Laboratory (PNL) has conducted two separate evaluations of the performance of extremity dosimeter processors to determine the appropriateness of the draft standard, as well as to obtain information regarding the performance of extremity dosimeters. Based on the information obtained during the facility visits and the results obtained from the performance testing, it was recommended that changes be made to ensure that the draft standard is appropriate for extremity dosimeters. The changes include: subdividing the mixture category and the beta particle category; eliminating the neutron category until appropriate flux-to-dose equivalent conversion factors are derived; and changing the tolerance level for the performance criterion to provide consistency with the performance criterion for whole body dosimeters, and to avoid making the draft standard overly difficult for processors of extremity dosimeters to pass. 20 refs., 10 figs., 6 tabs

  14. Electret dosimeter utilizing gas multiplication

    International Nuclear Information System (INIS)

    Ikeya, M.; Miki, T.

    1980-01-01

    It was found that the high electric field around the surface of an electret leads to cascade multiplication of the ionization process in a surrounding gas. Very sensitive charge decay constants of the order of 1mrad, were obtained for electrets composed of polyvinyliden fluoride or teflon polymers. The reduced charge is stable and can be utilized in personnel dosimetry. A simple pocket chamber dosimeter is described consisting of a small speaker or buzzer, a cylindrical chamber filled with air, argon or other gases, a polymer thermoelectret foil and an electrode. The sonic vibration of the foil induces an alternating charge on the electrode which is amplified and detected. The feasibility of this dosimeter and its shock and vibration resistance have been demonstrated. (author)

  15. Improving neutron dosimetry using bubble detector technology

    International Nuclear Information System (INIS)

    Buckner, M.A.

    1993-02-01

    Providing accurate neutron dosimetry for a variety of neutron energy spectra is a formidable task for any dosimetry system. Unless something is known about the neutron spectrum prior to processing the dosimeter, the calculated dose may vary greatly from that actually encountered; that is until now. The entrance of bubble detector technology into the field of neutron dosimetry has eliminated the necessity of having an a priori knowledge of the neutron energy spectra. Recently, a new approach in measuring personnel neutron dose equivalent was developed at Oak Ridge National Laboratory. By using bubble detectors in combination with current thermoluminescent dosimeters (TLDs) as a Combination Personnel Neutron Dosimeter (CPND), not only is it possible to provide accurate dose equivalent results, but a simple four-interval neutron energy spectrum is obtained as well. The components of the CPND are a Harshaw albedo TLD and two bubble detectors with theoretical energy thresholds of 100 key and 1500 keV. Presented are (1) a synoptic history surrounding emergence of bubble detector technology, (2) a brief overview of the current theory on mechanisms of interaction, (3) the data and analysis process involved in refining the response functions, (4) performance evaluation of the original CPND and a reevaluation of the same data under the modified method, (5) the procedure used to determine the reference values of component fluence and dose equivalent for field assessment, (6) analysis of the after-modification results, (7) a critique of some currently held assumptions, offering some alternative explanations, and (8) thoughts concerning potential applications and directions for future research

  16. Personnel neutron dosimetry applications of track-size distributions on electrochemically etched CR-39 foils

    International Nuclear Information System (INIS)

    Hankins, D.E.; Homann, S.G.; Westermark, J.

    1988-01-01

    The track-size distribution on electrochemically etched CR-39 foils can be used to obtain some limited information on the incident neutron spectra. Track-size distributions on CR-39 foils can also be used to determine if the tracks were caused by neutrons or if they are merely background tracks (which have a significantly different track-size distribution). Identifying and discarding the high-background foils reduces the number of foils that must be etched. This also lowers the detection limit of the dosimetry system. We have developed an image analyzer program that can more efficiently determine the track density and track-size distribution, as well as read the laser-cut identification numbers on each foil. This new image analyzer makes the routine application of track-size distributions on CR-39 foils feasible. 2 refs., 3 figs

  17. Personnel dosimetry of fast neutrons by silver activation in nuclear emulsions

    International Nuclear Information System (INIS)

    Francois, H.; Heilmann, C.; Jung, M.; Kappler, A.; Oppel, R.; Demoulin, R.

    1982-01-01

    This neutron dosimetry method may be extended to the radiological monitoring of a large number of workers. It uses photographic emulsion, a detector with long-established properties. The reproducibility of the detection characteristics is guaranteed by industrial manufacture. The method has been thoroughly tested and is now sufficiently under control for the application stage to be contemplated. The activation method has shown moreover that the optical counting technique accounts for only half the signal available. Owing to its sensitivity, energy response from 100 keV onwards, capacity to measure the neutron and electromagnetic (γ or X) doses simultaneously and complete automation the method may now be considered ready for extensive use in radioprotection [fr

  18. Remeasurement of early Harwell personnel film dosimeters

    International Nuclear Information System (INIS)

    Smith, J.W.; McGuinness, E.A.

    1981-05-01

    The A.E.A. epidemiological study depends for its radiation dose data on records of film badge doses which have been stored continuously since 1947. The records were originally kept to assist in the control of radiation hazard to the individual and for that purpose doses below the measurement threshold were unimportant. The epidemiological study could consider these doses to be either zero or equal to the measurement threshold, but both assumptions risk introducing a bias. Samples of old films worn in 1955 and 1960 were therefore remeasured using the present-day lower threshold of 0.005 rems to aid the choice of the most correct estimate. It was found that there had been little change in the blackening on the films or in the assessed radiation dose for 238 films above the original threshold of 0.05 rems, confirming that remeasurement was possible. The 2426 films previously recorded as ''<0.05 rems'' were found on remeasurement to give an average dose of 0.009 rems. A similar epidemiological study of radiation workers at Hanford took threshold doses as equal to zero and considered only positive measured doses. The errors possible with this approach are discussed. (author)

  19. Dosimeter charging apparatus

    International Nuclear Information System (INIS)

    Reuter, F.A.; Moorman, Ch.J.

    1985-01-01

    An apparatus for charging a dosimeter which has a capacitor connected between first and second electrodes and a movable electrode in a chamber electrically connected to the first electrode. The movable electrode deflects varying amounts depending upon the charge present on said capacitor. The charger apparatus includes first and second charger electrodes couplable to the first and second dosimeter electrodes. To charge the dosimeter, it is urged downwardly into a charging socket on the charger apparatus. The second dosimeter electrode, which is the dosimeter housing, is electrically coupled to the second charger electrode through a conductive ring which is urged upwardly by a spring. As the dosimeter is urged into the socket, the ring moves downwardly, in contact with the second charger electrode. As the dosimeter is further urged downwardly, the first dosimeter electrode and first charger electrode contact one another, and an insulator post carrying the first and second charger electrodes is urged downwardly. Downward movement of the post effects the application of a charging potential between the first and second charger electrodes. After the charging potential has been applied, the dosimeter is moved further into the charging socket against the force of a relatively heavy biasing spring until the dosimeter reaches a mechanical stop in the charging socket

  20. Application of solid dosimeter to radiation control

    International Nuclear Information System (INIS)

    Tsujimoto, Tadashi

    1988-01-01

    Individual exposure dose measuring devices are used to measure the dose of each person in facilities using radiations. Major devices of this type currently used in Japan include the film badge, thermoluminescence dosimeter, portable radiation dosimeter and fluorescent glass dosimeter. All of these devices except the portable radiation dosimeter are of a solid type. Various portable-type spatial dose rate measuring devices, generally called survey meters, are available to determine the spatial distribution of radiations. Major survey meters incorporates an ionization chamber, GM counter tube or scintillation counter, while BF 3 counting tubes are available for neutron measurement. Of these, the scintillation dosimeter is of a solid type. A new scintillation survey meter has recently been developed which incorporated a discrimination bias modulation circuit. Dosimeters incorporating an ionization chamber or a GM counter tube are generally used as portable alarms. Recently, a new solid-type alarm has been developed which incorporates a solicon radiation detector. Microcomputers are also used for self-diagnosis, data processing, automatic calibration, etc. (Nogami, K.)

  1. Personnel dose assignment practices

    International Nuclear Information System (INIS)

    Fix, J.J.

    1993-04-01

    Implementation of DOE N 5480.6 Radiological Control Manual Article 511(3) requirements, to minimize the assignment of personnel dosimeters, should be done only under a broader context ensuring that capabilities are in place to monitor and record personnel exposure both for compliance and for potential litigation. As noted in NCRP Report No. 114, personnel dosimetry programs are conducted to meet four major objectives: radiation safety program control and evaluation; regulatory compliance; epidemiological research; and litigation. A change to Article 511(3) is proposed that would require that minimizing the assignment of personnel dosimeters take place only following full evaluation of overall capabilities (e.g., access control, area dosimetry, etc.) to meet the NCRP objectives

  2. Use of CR-39 foils for personnel neutron dosimetry: improved electrochemical etching chambers and procedures

    International Nuclear Information System (INIS)

    Hankins, D.E.; Homann, S.G.; Westermark, J.

    1986-01-01

    The electrochemical etching procedures for the new dosimetry system that uses foils of CR-39 plastic has been improved. During 1985, the etching chambers were modified to correct several problems and the changes to the etching procedures were studied, which gave a more uniform track density and size. The currently recommended etch parameters are given. A new generation of CR-39 material from the manufacturer proved to have a considerably lower background track density and a higher sensitivity; the new foils are also more uniform in thickness, which eliminates the need to numerically compensate for thickness variations. The energy dependence of the CR-39 using monoenergetic neutrons from accelerators at Battelle Northwest Laboratories and at Los Alamos National Laboratory was determined. Some variation was found in the energy dependence, but it is believed this was caused by changes in the etching procedures and by uncertainties in the fluences of the neutrons from the accelerators. A means by which the counting of CR-39 tracks may be automated is suggested; this would be very useful in adapting the CR-39 dosimetry system to large-scale use

  3. Report on the Personnel Dosimetry at AB Atomenergi during 1964

    International Nuclear Information System (INIS)

    Edvardsson, K.A.; Hagsgaard, S.

    1966-01-01

    This report presents the results of the personnel dosimetry at AB Atomenergi during 1964. No doses exceeding the recommendations of ICRP were reported. The sum of the reported external total body doses during the year was for AB Atomenergi 51.5 manrem which, distributed over the whole company personnel, corresponds to an average dose of about 35 mrem per year and person or less than 1 % of the maximum permissible dose. 31,400 gamma films and 5,800 neutron films were evaluated. The films were changed every month. Urine analyses numbered 2,731 and whole body measurements 485. A comparison is made between dose distributions at AB Atomenergi and at institutions in other countries. The fraction of all personnel carrying dosimeters and exposed to more than a nominal dose seems generally to have been less than 10-20 %

  4. Report on the Personnel Dosimetry at AB Atomenergi during 1964

    Energy Technology Data Exchange (ETDEWEB)

    Edvardsson, K A; Hagsgaard, S

    1966-01-15

    This report presents the results of the personnel dosimetry at AB Atomenergi during 1964. No doses exceeding the recommendations of ICRP were reported. The sum of the reported external total body doses during the year was for AB Atomenergi 51.5 manrem which, distributed over the whole company personnel, corresponds to an average dose of about 35 mrem per year and person or less than 1 % of the maximum permissible dose. 31,400 gamma films and 5,800 neutron films were evaluated. The films were changed every month. Urine analyses numbered 2,731 and whole body measurements 485. A comparison is made between dose distributions at AB Atomenergi and at institutions in other countries. The fraction of all personnel carrying dosimeters and exposed to more than a nominal dose seems generally to have been less than 10-20 %.

  5. Personnel radiation dosimetry

    International Nuclear Information System (INIS)

    1987-01-01

    The book contains the 21 technical papers presented at the Technical Committee Meeting to Elaborate Procedures and Data for the Intercomparison of Personnel Dosimeters organizaed by the IAEA on 22-26 April 1985. A separate abstract was prepared for each of these papers. A list of areas in which additional research and development work is needed and recommendations for an IAEA-sponsored intercomparison program on personnel dosimetry is also included

  6. Dosimeter design specifications

    International Nuclear Information System (INIS)

    Anon.

    1981-01-01

    The combination dosimeter and security credential holder was developed as part of the effort involved to provide an automated readout and thermoluminescent dosimetry capability at Hanford. The holder is designed to accomodate the thermoluminescent dosimeter card, appropriate filters, the security credential and a snap type clip. The body of the holder is ABS plastic (acrylontrile-butadiene-styrene). The dosimeter holder and card is mold casted providing uniformity of construction

  7. Thermoluminescent dosimeter system

    International Nuclear Information System (INIS)

    Felice, P.E.; Gonzalez, J.L.; Seidel, J.G.

    1979-01-01

    An improved thermoluminescent dosimeter system and apparatus for sensing alpha particle emission is described. A thermoluminescent body is sealed between a pair of metallized plastic films. The dosimeter is mounted within a protective inverted cup or a tube closed at one end, which is disposed in a test hole for exposure to radioactive radon gas which is indicaive of uranium deposits

  8. SDI-100 radiation dosimeter

    International Nuclear Information System (INIS)

    Zheng Zheng; Zhao Yongfu; Dai Honggui

    1995-01-01

    An intelligent radiation dosimeter, with such functions as signal collection and data processing, store, print and display, has been developed. Its detector is made of a micro-semiconductor. This dosimeter can be used in laboratories for agricultural 60 Co irradiators, radiotherapeutic facilities and other small and medium-size 60 Co irradiators

  9. Neutron dosimetry program at Mound - problems and solutions

    International Nuclear Information System (INIS)

    Winegardner, M.K.

    1991-01-01

    The Mound personnel neutron dosimetry program utilizes TLD albedo technology. The neutron dosimeter design incorporates a two-element spectrometer for site-specific neutron quality determination and empirical application of field neutron calibration factors. Design elements feature two Li(6)F (TLD- 600) chips for neutron detection and one Li(7)F (TLD-700) chip for gamma compensation of the TLD- 600 chips. One TLD-600 chip is Cadmium shielded on the front side of the dosimeter, the other is Cadmium shielded from the back side. Tin filters are placed opposite of the Cadmium shield on each of the TLD-600 chips and on both sides of the TLD-700 chip for symmetrically equivalent gamma absorption characteristics. Neutron quality determination is accomplished by the albedo neutron-to- incident thermal neutron response ratio above the Cadmium cutoff. This front Cadmium shielded-to-back Cadmium shielded response ratio, compensated for the presence of gamma radiation, provides the basis for neutron energy calibration via the albedo response curve

  10. Passive radon daughter dosimeters

    International Nuclear Information System (INIS)

    McElroy, R.G.C.; Johnson, J.R.

    1986-03-01

    On the basis of an extensive review of the recent literature concerning passive radon daughter dosimeters, we have reached the following conclusions: 1) Passive dosimeters for measuring radon are available and reliable. 2) There does not presently exist an acceptable passive dosimeter for radon daughters. There is little if any hope for the development of such a device in the foreseeable future. 3) We are pessimistic about the potential of 'semi-passive dosimeters' but are less firm about stating categorically that these devices cannot be developed into a useful radon daughter dosimeter. This report documents and justifies these conclusions. It does not address the question of the worker's acceptance of these devices because at the present time, no device is sufficiently advanced for this question to be meaningful. 118 refs

  11. Modern methods of personnel dosimetry

    International Nuclear Information System (INIS)

    Kraus, W.; Herrmann, D.; Kiesewetter, W.

    The physical properties of radiation detectors for personnel dosimetry are described and compared. The suitability of different types of dosimeters for operational and central monitoring of normal occupational exposure, for accident and catastrophe dosimetry and for background and space-flight dosimetry is discussed. The difficulties in interpreting the dosimeter reading with respect to the dose in individual body organs are discussed briefly. 430 literature citations (up to Spring 1966) are given

  12. Alarm pocket dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Hiraki, H; Kitamura, S [Matsushita Electric Industrial Co. Ltd., Kadoma, Osaka (Japan)

    1975-04-01

    This instrument is a highly reliable pocket dosimeter which has been developed for personal monitoring use. The dosimeter generates an alarm sound when the exposure dose reaches a preset value. Using a tiny GM tube for a radiation detector and measuring the integrated dose by means of a digital counting method, this new pocket dosimeter has high accuracy and stability. Using a sealed alkali storage battery for the power supply, and with an automatic control charger, this dosimetry system is easy and economical to operate and maintain. Detectable radiation by the dosimeter are X and ..gamma.. rays. Standard preset dose values are 30, 50, 80 and 100 mR. Detection accuracy is betwen +10% and -20%. The dosimeter is continuously usable for more than 14 hours after charging for 2 hours. The dosimeter has the following features; good realiability, shock-proof loud and clear alarm sound, the battery charger also serves as a stock container for the dosimeters, and no switching operation required for the power supply due to the internal automatic switch. Therefore, the dosimetry system is very useful for personal monitoring management in many radiation industry establishments.

  13. 10 CFR 36.55 - Personnel monitoring.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 1 2010-01-01 2010-01-01 false Personnel monitoring. 36.55 Section 36.55 Energy NUCLEAR REGULATORY COMMISSION LICENSES AND RADIATION SAFETY REQUIREMENTS FOR IRRADIATORS Operation of Irradiators § 36.55 Personnel monitoring. (a) Irradiator operators shall wear a personnel dosimeter that is...

  14. Evaluation of personal dosimeters

    International Nuclear Information System (INIS)

    Correa, C. A.

    2007-01-01

    This work makes a screening of the different types of dosimeters present in the international market, to provide operative dosimetry of individual monitoring to measure Hp(10) and Hp(0,07)-specifically for external radiation gamma and beta, as well as to give knowledge of advances of passive and operative dosimetry, and the changes in the regulatory policy relative to these aspects. The data has been extracted from several providers of dosimeters, and the importance has been stressed in a good election of the dosimeter before its use, as well as the important advances in these equipment. (Author) 14 refs

  15. Electrochemical development of particle tracks in CR-39 polymer dosimeter

    International Nuclear Information System (INIS)

    Hadlock, D.E.; Parkhurst, M.A.; Yang, C.S.; Groeger, J.; Johnson, J.R.; Huang, S.J.

    1985-09-01

    Electrochemical etching of CR-39 polymeric track etch neutron detectors results in proton-recoil tracks can be distinguished from background tracks much better than tracks developed solely by chemical etching. A newly designed and constructed electrochemical etching apparatus allows large numbers of dosimeters to be processed simultaneously with consistent results. Many processing systems have been developed for chemical and electrochemical etching of the track etch dosimeters. Three systems specifically show great promise and are being studied extensively

  16. DEPRON dosimeter for ``Lomonosov'' satellite

    Science.gov (United States)

    Brilkov, Ivan; Vedenkin, Nikolay; Panasyuk, Mikhail; Amelyushkin, Aleksandr; Petrov, Vasily; Nechayev, Oleg; Benghin, Victor

    appearance of the instrument DEPRON (Dosimeter of Electrons, PROtons and Neutrons) was determined. DEPRON is intended for registration of the absorbed doses and linear energy transfer spectra for high-energy electrons, protons and nuclei of space radiation, as well as registration of thermal and slow neutrons. The experiment based on DEPRON instrument is aimed at the studies of the distribution of space radiation dose rate at high latitude paths in order to study the flight paths of perspective manned spacecraft. Present work provides a brief description of the DEPRON instrument, its calibration results and the structure of the output data.

  17. Automatic dosimeter for kerma measurement based on commercial PIN photo diodes

    International Nuclear Information System (INIS)

    Kushpil, V.; Kushpil, S.; Huna, Z.

    2011-01-01

    A new automatic dosimeter for measurement of radiation dose from neutron and ionization radiation is presented. The dosimeter (kerma meter) uses commercial PIN diodes with long base as its active element. Later it provides a maximal dependence of the minority carriers life time versus absorbed dose. The characteristics of the dosimeter were measured for several types of commercial diodes. Device can be useful in many environmental or industrial applications. (authors)

  18. Limitations of commonly used thick-element personal dosimeters

    International Nuclear Information System (INIS)

    Gupta, V.P.

    1983-01-01

    In the ANSI Standard N13.11, accepted in June 1982, radiation dose depths of 1.0 cm and 0.007 cm in tissue for protection dosimetry have been adopted for deep and shallow dose equivalent estimations respectively. This standard is presently used for a mandatory personnel dosimetry performance testing program in the United States. Estimation of shallow-dose equivalent using a two-element dosimeter is described under the guidelines of this standard and the dosimetry practices followed by most dosimeter processors. A mathematical formulation, correlating a dosimeter response and shallow-dose equivalent factors at different energies, is presented. Also, the performance of a two-element thermoluminescent dosimeter is examined and the shallow-dose equivalent response results, both for the beta particles and photons, are discussed

  19. Radon daughter dosimeter

    International Nuclear Information System (INIS)

    Durkin, J.

    1977-01-01

    This patent describes a portable radon daughter dosimeter unit used to measure radon gas alpha daughters in ambient air. These measurements can then be related to preselected preestablished standards contained in a remote central readout unit. The dosimeter unit is adapted to be worn by an operator in areas having alpha particle radiation such as in uranium mines. Within the dosimeter is a detector head housing having a filter head and a solid state surface barrier radiation detector; an air pump to get air to the detector head; a self contained portable power supply for the unit; and electronic circuitry to process detected charged electrons from the detector head to convert and count their pulses representatives of two alpha radon emitter daughters. These counted pulses are in binary form and are sent to a readout unit where a numerical readout displays the result in terms of working level-hours

  20. Radon daughter dosimeter

    International Nuclear Information System (INIS)

    Durkin, J.

    1977-01-01

    A portable radon daughter dosimeter unit used to measure Radon gas alpha daughters in ambient air is described. These measurements can then be related to preselected preestablished standards contained in a remote central readout unit. The dosimeter unit is adapted to be worn by an operator in areas having alpha particle radiation such as uranium mines. Within the dosimeter is a detector head housing having a filter head and a solid state surface barrier radiation detector; an air pump to get air to the detector head; a self contained portable power supply for the unit; and electronic circuitry to process detected charged electrons from the detector head to convert and count their pulses representatives of two alpha radon emitter daughters. These counted pulses are in binary form and are sent to a readout unit where a numerical readout diplays the result in terms of working level-hours

  1. Fundamentals of Polymer Gel Dosimeters

    Science.gov (United States)

    McAuley, Kim B.

    2006-12-01

    The recent literature on polymer gel dosimetry contains application papers and basic experimental studies involving polymethacrylic-acid-based and polyacrylamide-based gel dosimeters. The basic studies assess the relative merits of these two most commonly used dosimeters, and explore the effects of tetrakis hydroxymethyl phosphonium chloride (THPC) antioxidant on dosimeter performance. Polymer gel dosimeters that contain THPC or other oxygen scavengers are called normoxic dosimeters, because they can be prepared under normal atmospheric conditions, rather than in a glove box that excludes oxygen. In this review, an effort is made to explain some of the underlying chemical phenomena that affect dosimeter performance using THPC, and that lead to differences in behaviour between dosimeters made using the two types of monomer systems. Progress on the development of new more effective and less toxic dosimeters is also reported.

  2. Calibration of an ALBEDO termoluminiscent dosimeter for its use in personal dosimetry

    International Nuclear Information System (INIS)

    Diaz Bernal, E.; Molina Perez, D.; Cornejo Diaz, N.; Carrazana Gonzalez, J.

    1996-01-01

    The dosimetric studies began after the Radiological Individuals Surveillance Department from the Radiation Protection and Hygiene Center acquired the albedo thermoluminescent dosimeters model JR1104. This paper reviews the response of those dosimeters to the different spectrums and incidence angles of neutronic radiation

  3. Radiation dosimeter assembly

    International Nuclear Information System (INIS)

    Seidel, J.G.

    1982-01-01

    A technique is disclosed for securing a thermoluminescent radiation dosimeter, used for monitoring underground radon gas in uranium prospecting, to a cup-like support member made of heavy gauge aluminum foil. A metalized film, consisting of an aluminum layer and a high tensile strength plastic layer, covers an aperture in the support members for the dosimeter. The film is secured by a high temperature adhesive to the support member, and both are capable of withstanding an annealing temperature of up to 300 0 C

  4. Study of personnel monitoring programs

    International Nuclear Information System (INIS)

    Guerra, Aline B.; Lorenzini, Fabiane; Carlos, Janaina; Bernasiuk, Maria E.B.; Rizzatti, Mara R.; Fuentefria, Jose L.B.

    1996-01-01

    Surveillance of several health institutions who use ionizing radiation sources, as well as data from the state of Rio Grande do Sul (Brazil) are studied concerning the use of personnel dosimeters. The results show that several institutions do not provide them and those which provide do not know how to use them

  5. Human hair as a pollutant dosimeter

    International Nuclear Information System (INIS)

    Al-Hashimi, A.

    1991-01-01

    Human hair has been proved to be a better dosimeter than even blood for tracing most of the heavy metal toxins when they penetrate the biosphere. The high precision of the neutron activation analysis (NAA) enabled researchers to elegantly differentiate between endogenous and exogenous contamination and thoroughly study poisonings caused by these physiologically-unimportant elements. Extensive volume of bench-scale work has been accomplished in these laboratories to show the capacity of INAA to detect the presence of 10 nuclides (or more) with a precision of about 5%. The principal objective of the present study is to employ this assaying power and the tendency of scalp hair to uptake heavy metals from aqueous solutions, to design a dosimeter which can easily be used by the environmentalists. The findings should also be of interest to the waste-management people who are searching for a cost-effective technique to remove these pollutants from relatively large volumes of industrial effluents

  6. Thermoluminescence dosimeter reader

    International Nuclear Information System (INIS)

    Robertson, M.E.A.; Marshall, J.; Brabants, J.A.P.; Davies, M.E.

    1975-01-01

    An electric circuit arrangement is described including a photomultiplier tube and a high voltage source therefor also includes a feedback loop from the output of the tube to the high voltage source, and loop providing automatic gain stabilization for the tube. The arrangement is used in a dosimeter reader to provide sensitivity correction for the reader each time the reader is to be used

  7. Evaluating the BD-100R [Bubble Technology Industries] as a neutron spectrometer through pressure variation

    International Nuclear Information System (INIS)

    White, B.; Ebert, D.; Munno, F.

    1990-01-01

    Other investigators have evaluated the dose response of neutron bubble dosimeters for possible use as personnel monitors for the U.S. Navy in low level radiation fields. In addition to dose measurements, these devices have been modified to measure the neutron energy spectra. These methods tend to be cumbersome, inaccurate, or both and do not use the same devices as employed in the dosimetry. The BD-100R dosimeter used in this work consists of a test tube containing an elastic polymer with interspersed droplets of two types of Freon; Freon-12 and Freon-114. Each superheated liquid droplet is a potential nucleation site. The minimum energy needed to form a bubble from the nucleation site is given by, E = 16πγ 3 (T)/3(ΔP) 2 , where ΔP is the difference between the vapor pressure of the droplet and the applied pressure. Upon reaching a critical radius, a bubble becomes unstable and grows in size. It may be seen from this equation that the energy deposition required for bubble formation is inversely proportional to the square of the pressure difference. The number of bubbles formed continually decreases with increasing applied pressure, until a pressure is reached where no bubbles are formed since the energy transferred can no longer vaporize the Freon. This investigation is intended to demonstrate the determination of an unknown spectrum utilizing the dosimeter response (number of bubbles formed) as a function of the neutron energy (applied pressure). A set of 12 dosimeters was initially exposed outside the East Beam Port (EBP) at the Maryland University Training Reactor (MUTR). The dosimeters were inside a pressure chamber which could accommodate up to 18 dosimeters. The same set of dosimeters were irradiated using a paraffin moderated PuBe source for which the neutron energy spectrum is unknown. There were eight exposures of six dosimeters at varied pressures in the EBP. The average number of bubbles and standard deviation was measured for each pressure. Data

  8. Personnel dosimetry in fluoroscopy

    International Nuclear Information System (INIS)

    Baechler, S.; Gardon, M.; Bochud, F.; Sans-Merce, M.; Verdun, F.R.; Trueb, Ph.

    2006-01-01

    Physicians who frequently perform fluoroscopic examinations are exposed to high intensity radiation fields and should use protective equipment such as lead aprons, thyroid shields and lead glasses. Standard individual dosimeters are worn under the lead apron in order to measure a dose that is representative of effective dose. However, large parts of the body are not protected by the apron (e.g. arms, head). Given a protection factor for the apron of about 100, an important irradiation of a body part not under the apron could go undetected. A study was conducted to analyse this situation by measuring dose using two dosimeters, one over-apron and one under-apron, for radiologists performing frequent fluoroscopic examinations. Measurements made over six-month period show that, indeed, the use of a single under-apron dosimeter is inadequate for personnel monitoring. Large doses to the head and arms are going undetected by this technique. A method for weighting the doses measured by under- and over-apron dosimeters to obtain a value better representative of the effective dose will be proposed. (authors)

  9. PNNL Results from 2010 CALIBAN Criticality Accident Dosimeter Intercomparison Exercise

    International Nuclear Information System (INIS)

    Hill, Robin L.; Conrady, Matthew M.

    2011-01-01

    This document reports the results of the Hanford personnel nuclear accident dosimeter (PNAD) and fixed nuclear accident dosimeter (FNAD) during a criticality accident dosimeter intercomparison exercise at the CEA Valduc Center on September 20-23, 2010. Pacific Northwest National Laboratory (PNNL) participated in a criticality accident dosimeter intercomparison exercise at the Commissariat a Energie Atomique (CEA) Valduc Center near Dijon, France on September 20-23, 2010. The intercomparison exercise was funded by the U.S. Department of Energy, Nuclear Criticality Safety Program, with Lawrence Livermore National Laboratory as the lead Laboratory. PNNL was one of six invited DOE Laboratory participants. The other participating Laboratories were: Lawrence Livermore National Laboratory (LLNL), Los Alamos National Laboratory (LANL), Savannah River Site (SRS), the Y-12 National Security Complex at Oak Ridge, and Sandia National Laboratory (SNL). The goals of PNNL's participation in the intercomparison exercise were to test and validate the procedures and algorithm currently used for the Hanford personnel nuclear accident dosimeters (PNADs) on the metallic reactor, CALIBAN, to test exposures to PNADs from the side and from behind a phantom, and to test PNADs that were taken from a historical batch of Hanford PNADs that had varying degrees of degradation of the bare indium foil. Similar testing of the PNADs was done on the Valduc SILENE test reactor in 2009 (Hill and Conrady, 2010). The CALIBAN results are reported here.

  10. Evaluation of discrepancies between thermoluminescent dosimeter and direct-reading dosimeter results

    International Nuclear Information System (INIS)

    Shaw, K.R.

    1993-07-01

    Currently at Oak Ridge National Laboratory (ORNL), the responses of thermoluminescent dosimeters (TLDs) and direct-reading dosimeters (DRDs) are not officially compared or the discrepancies investigated. However, both may soon be required due to the new US Department of Energy (DOE) Radiological Control Manual. In the past, unofficial comparisons of the two dosimeters have led to discrepancies of up to 200%. This work was conducted to determine the reasons behind such discrepancies. For tests conducted with the TLDs, the reported dose was most often lower than the delivered dose, while DRDs most often responded higher than the delivered dose. Trends were identified in personnel DRD readings, and ft was concluded that more training and more control of the DRDs could improve their response. TLD responses have already begun to be improved; a new background subtraction method was implemented in April 1993, and a new dose algorithm is being considered. It was concluded that the DOE Radiological Control Manual requirements are reasonable for identifying discrepancies between dosimeter types, and more stringent administrative limits might even be considered

  11. Performance evaluation of a colorimetric hydrazine dosimeter

    Science.gov (United States)

    Brenner, Karen P.; Rose-Pehrsson, Susan L.

    1994-06-01

    A dosimeter for real-time, colorimetric detection of hydrazine in air has been developed. The passive badge consists of a dosimeter card containing a vanillin solution coated on a thin paper substrate. The active patch consists of a thick cellulose substrate coated with a vanillin solution. When placed in a plastic sample holder attached to a personnel pump, up to 5 L/min can be drawn through the active badge substrate. Through a condensation reaction, vanillin reacts with hydrazine to form a colored product that absorbs in the visible region. The hydrazone formed in the reaction is yellow; its intensity is proportional to the dose. When exposed passively to hydrazine, the experimental detection limit is less than 20 ppb-hrs. Extrapolated results indicate a detection limit of less than 5 ppb-hrs for long sampling periods. Actively sampling of hydrazine vapors gives an experimental detection limit of less than 100 ppb-L at a sample rate of 5 L/min. Relative humidity effects on badge response were minor. High humidity enhanced the color development on the vanillin badge; while low humidity had no effect on badge response. Interference testing of the dosimeters revealed a tobacco smoke interference. Preliminary shelf life tests indicated no decrease in sensitivity to hydrazine when stored at room temperature for 6 months.

  12. Evaluation of optical fibres as gamma radiation dosimeter

    International Nuclear Information System (INIS)

    Bohra, Dinesh; Chaudhary, H.S.; Panwar, Lalit; Vaijapurkar, S.G.; Bhatnagar, P.K.; Dasgupta, K.

    2005-01-01

    Semiconductor base gamma and neutron sensors are the fastest and popular dosimeters and are in competition with Thermoluminescence (TL) and Radio photoluminescence (RPL) dosimeters. All over the world armed forces require a dosimeter which records cumulative doses of ionizing radiations from mcGy to 10 Gy and is readable repeatedly without loss of dose information. TL dosimeters do not meet the criteria and RPL dosimeter meet the expectations and are in use by armed forces. Technologists have used laser as an excitation source to stimulate the glass and have achieved success in recording gamma doses of occupational/accidental span (mcGy to 10 Gy). However synthesizing RPL glass batches with exactly same characteristics predoses is a difficult task. Silicon base phosphorous doped step index multimode optical fibre can be made in a significant quantity and large number of dosimeters from it can be achieved with uniform predose. The radiation induced transmission loss gives a measure of gamma dose which is cumulative, readable repeatedly without loss of information. Assorted composition, core dia optical fibres have been synthesized and evaluated for dose linearity, dose rate independence, fading, length optimization. Here in is described some results of recent experiments and sensitivities achieved. (author)

  13. The neutron and low-energy gamma operational dosimetry in Melox plant

    International Nuclear Information System (INIS)

    Devita, A.D.

    2006-01-01

    M.E.L.O.X., subsidiary of A.R.E.V.A., produce M.O.X. fuels, a mixture of uranium and plutonium oxides. With the use in the process of plutonium oxide, there is a risk of external exposure to neutrons and low -energy gamma rays. By their characteristics, both these types of radiation are difficult to measure. The difficulty in measuring neutron doses lies in the fact that the fluence -to-dose equivalent conversion factor varies with the neutron energy level. In low -energy gamma (between 20 and 60 keV) dose measurement, the problem is detection using an electronic system. Just some years ago, very few industrial players were tempted to develop dosimeters in these areas in view of the poor demand and market prospects. Furthermore, radiation protection specialists needed a highly functional and robust direct reading dosimeters or, in other words, a device that was simple, reliable, inexpensive, small, and quick and easy to use in a wide range of working environments that could vary in terms of both the workstation and external exposure. In addition, at sites such as Melox, where company employees work alongside personnel from outside companies, the same types of dosimeters must be used so that dose -related data can be managed globally in one data base. Two technical solutions are available for neutron operational dosimetry - spectrometer-dosimeters and calibration dosimeters. Melox has opted for the use of calibration dosimeters. The reasons for this choice (technical, financial and organizational criteria) are given in this presentation. Before and during the various campaigns of M.O.X. fuels, the spectral characteristics relating to neutron fluence at different workstations and representative of personnel exposure levels were determined. A reference spectrometer was then used to determine the transfer function between fluence and dose in order to calibrate passive and operational dosimeters appropriately.The methodology to be set up should guarantee good

  14. Personnel radiation dose assessment using a novel dosimeter in the department of radiology and dentistry in a medical facility in Delta State, South-South Nigeria: Our experience in the last 4 years

    OpenAIRE

    Omojola, Akintayo Daniel; Akpochafor, Michael Onoride; Adeneye, Samuel Olaolu; Aniekop, Ukeme Pius; Anizor, Margaret Idongesit; Ekpo, Mary-Ann Etim; Madu, Chibuzor Bede

    2018-01-01

    Objectives. Large percentagesof X-ray facilities in Nigeria do not use radiation monitoring device; a fewpercentage that use them do not evaluate or carryout out assessment programs toascertain the detriment to occupationally exposed workers. This study was aimedat evaluating dose reports from 2013 to 2016 for personnel who operate  radiation facilities and those that workwithin radiation field during certain X-ray procedures/examinations in thedepartment of radiology and dentistry respective...

  15. X-rays individual dose assessment using TLD dosimeters

    International Nuclear Information System (INIS)

    Salas, Carlos

    2008-01-01

    This paper describes the methodology used in Embalse NPP for measuring individual X-ray dose in dentists and radiologists, who work in areas near the plant. Personnel is provided with TLD personal dosimeters for thoracic use, as well as TLD ring dosimeters. This individual X-ray dosimetry is fundamental in order to know the effective energy coming from the radiation field, since the dosimetry factors depend on it. On the other hand, the response of the TLD crystals also depends of the effective energy; this accentuates the problem when assessing the individual dose. The X-ray dosimeter must simultaneously determine the value of the effective energy and the corresponding dose value. The basic principle for determining effective energy is by using at least two different TLD materials covered by filters of different thickness. The TLD materials used have totally energy responses. Therefore, different readouts from each of the crystals are obtained. The ratio between both readouts provides a factor that depends of the effective energy but that is 'independent' from the exposure values irradiated to the dosimeter. The Personal TLD dosimeter currently in use is Bicron-Harshaw. It comprises a carrier model 8807. This carrier contains a card model 2211 which groups two TLD 200 crystals and two TLD 100 crystals. It has internal filters at each side of the TLD 200 crystals. The periodical calibration of these dosimeters consists in the irradiation of some dosimeters with different X-ray energy beams in the National Atomic Energy Commission (CNEA). This dosimeter was used, by the National Regulatory Authority (ARN) in several comparisons, always getting satisfactory results. (author)

  16. CRRES dosimeter simulations

    International Nuclear Information System (INIS)

    Auchampaugh, G.; Cayton, T.

    1993-04-01

    Conflicting data have been obtained from electron instruments aboard CRRES (Combined Release and Radiation Effects Satellite). To gain insight and to help in the interpretation of the data, we have calculated electron- and proton-flux and dose response functions for the four domes of the CRRES dosimeters using the Los Alamos Monte Carlo radiation transport codes. The response functions were calculated for electron and proton energies representative of those present in the space radiation environment. We also calculated the response of the dosimeters to a model radiation environment for orbit 607, which occurred on April 1, 1991 and compared the results to the measured values. The electron and proton components of the radiation environment were calculated using the solar maximum versions of the AE8 and AP8 models, namely, AE8MAX and AP8MAX. To facilitate the second task, we wrote two FORTRAN programs (CRRESunderscoreSIMP for AP8MAX and CRRESunderscoreSIME for AE8MAX) to read in a standard CRRES data file and to produce a comparison file of the calculated and measured values for all four dosimeter domes.The FORTRAN code will be available to the Phillips Laboratory for their use in making comparisons to other orbital data

  17. Radiation dosimeters for medical use

    International Nuclear Information System (INIS)

    Risticj, S. Goran

    2013-01-01

    The several personal radiation dosimeter types for medical use, which look like promising for this kind of application, as pMOS (RADFET) dosimeter, direct ion storage (DIS) dosimeters, thermoluminescent (TL) and optically stimulated luminescent (OSL) dosimeters, are described, and their advantages and disadvantages are analyzed. The p-channel metal-oxide-semiconductor (pMOS) dosimetric transistors allow dose measurements in vivo in real time, and they are especially important for radiotherapy. Direct ion storage (DIS) dosimeters are a hybrid of ion chamber and floating gate MOSFETs (FGMOSFETs), show very high sensitivity. Radiative processes that happen during the exposure of crystal to radiation are classified as prompt luminescence or radioluminescence (RL). In the case of an emission during stimulation, this phenomenon is referred to thermoluminescence or optically stimulated luminescence depending on whether the stimulation source is heat or light. TL and OSL dosimeters are natural or synthetic materials, which the intensity of emitted light is proportional to the irradiation dose. (Author)

  18. Technical guidelines for personnel dosimetry calibrations

    International Nuclear Information System (INIS)

    Roberson, P.L.; Fox, R.A.; Hadley, R.T.; Holbrook, K.L.; Hooker, C.D.; McDonald, J.C.

    1983-01-01

    A base of technical information has been acquire and used to evaluate the calibration, design, and performance of selected personnel systems in use at Department of Energy (DOE) facilites. A technical document was prepared to guide DOE and DOE contractors in selecting and evaluating personnel dosimetry systems and calibration. A parallel effort was initiated to intercompare the adiological calibrations standards used to calibrate DOE personnel dosimeters

  19. Copper doped borate dosimeters revisited

    Energy Technology Data Exchange (ETDEWEB)

    Alajerami, Y.S.M. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Department of Medical Radiography, Al-Azhar University, Gaza Strip, Palestine (Country Unknown); Hashim, S., E-mail: suhairul@utm.my [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Oncology Treatment Centre, Sultan Ismail Hospital, 81100 Johor Bahru (Malaysia); Ghoshal, S.K. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Bradley, D.A. [Centre for Nuclear and Radiation Physics, Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Department of Physics, Faculty of Science, University of Malaya, 50603 Kuala Lumpur (Malaysia); Mhareb, M. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Saleh, M.A. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); National Atomic Energy Commission (NATEC), Sana' a (Yemen)

    2014-11-15

    We render a panoramic overview on copper (Cu) doped borate dosimeters. Preparing a dosimeter by mixing specific materials with precise weights and methods is a never-ending quest. The recommended composition is highly decisive for accurate estimation of the absorbed dose, prediction of the biological outcome, determination of the treatment dose for radiation therapy and facilitation of personal monitoring. Based on these principles, the proposed dosimeter must cover a series of dosimetric properties to realize the exact results and assessment. The doped borate dosimeters indeed demonstrate attractive thermoluminescence (TL) features. Several dedicated efforts are attempted to improve the luminescence properties by doping various transition metals or rare-earth elements. The Cu ion being one of the preferred activators shows excellent TL properties as revealed via detail comparison with other dosimeters. Two oxide states of Cu (Cu{sup +} and Cu{sup ++}) with reasonable atomic number allow easy interaction with boron network. Interestingly, the intrinsic luminescent centers of borate lattice are in cross linked with that of Cu{sup +} ions. Thus, the activation of borate dosimeter with Cu ions for the enhancement of the TL sensitivity is recognized. These dosimeters reveal similar glow curves as the standard TLD-100 (LiF:Mg,Ti) one irrespective of the use of modifiers and synthesis techniques. They display high sensitivity, low fading, dose response linearity over wide range and practical minimum detectable dose. Furthermore, the effective atomic number being the most beneficial aspect (equivalent to that of human tissue) of borate dosimeters do not show any change due to Cu ion activations. The past development, major challenges, excitement, applications, recent progress and the future promises of Cu doped borate TL dosimeters are highlighted. - Highlights: • The manuscript gives a panoramic overview on copper doped borate dosimeters. • Cu ions activated

  20. Copper doped borate dosimeters revisited

    International Nuclear Information System (INIS)

    Alajerami, Y.S.M.; Hashim, S.; Ghoshal, S.K.; Bradley, D.A.; Mhareb, M.; Saleh, M.A.

    2014-01-01

    We render a panoramic overview on copper (Cu) doped borate dosimeters. Preparing a dosimeter by mixing specific materials with precise weights and methods is a never-ending quest. The recommended composition is highly decisive for accurate estimation of the absorbed dose, prediction of the biological outcome, determination of the treatment dose for radiation therapy and facilitation of personal monitoring. Based on these principles, the proposed dosimeter must cover a series of dosimetric properties to realize the exact results and assessment. The doped borate dosimeters indeed demonstrate attractive thermoluminescence (TL) features. Several dedicated efforts are attempted to improve the luminescence properties by doping various transition metals or rare-earth elements. The Cu ion being one of the preferred activators shows excellent TL properties as revealed via detail comparison with other dosimeters. Two oxide states of Cu (Cu + and Cu ++ ) with reasonable atomic number allow easy interaction with boron network. Interestingly, the intrinsic luminescent centers of borate lattice are in cross linked with that of Cu + ions. Thus, the activation of borate dosimeter with Cu ions for the enhancement of the TL sensitivity is recognized. These dosimeters reveal similar glow curves as the standard TLD-100 (LiF:Mg,Ti) one irrespective of the use of modifiers and synthesis techniques. They display high sensitivity, low fading, dose response linearity over wide range and practical minimum detectable dose. Furthermore, the effective atomic number being the most beneficial aspect (equivalent to that of human tissue) of borate dosimeters do not show any change due to Cu ion activations. The past development, major challenges, excitement, applications, recent progress and the future promises of Cu doped borate TL dosimeters are highlighted. - Highlights: • The manuscript gives a panoramic overview on copper doped borate dosimeters. • Cu ions activated technique in borate

  1. Nevada test site neutron dosimetry-problems/solutions

    International Nuclear Information System (INIS)

    Sygitowicz, L.S.; Bastian, C.T.; Wells, I.J.; Koch, P.N.

    1991-01-01

    Historically, neutron dosimetry at the NTS was done using NTA film and albedo LiF TLD's. In 1987 the dosimeter type was changed from the albedo TLD based system to a CR-39 track etch based system modeled after the program developed by D. Hankins at LLNL. Routine issue and return is performed quarterly for selected personnel using bar-code readers at permanent locations. The capability exists for work site issue as-needed. Issue data are transmitted by telephone to a central computer where it is stored until the dosimeter is returned, processed and read, and the dose calculation is performed. Dose equivalent calculations are performed using LOTUS 123 and the results are printed as a hard copy record. The issue and dose information are hand-entered into the Dosimetry database. An application is currently being developed to automate this sequence

  2. Characteristics of radiophotoluminescent glass dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Ito, Masashi; Shiraishi, Akemi; Murakami, Hiroyuki [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    2001-07-01

    In Japan Atomic Energy Research Institute, a film badge is recently replaced by a new type radiophotoluminescent (RPL) glass dosimeter for external personal monitoring. Some fundamental characteristics of this dosimeter, such as dose dependence linearity, energy dependence, angular dependence, dose evaluation accuracy at mixed irradiation conditions, fading, etc., were examined at the Facility of Radiation Standard (FRS), JAERI. The results have proved that the RPL glass dosimeter has sufficient characteristics for practical use as a personal dosimeter for all of the items examined. (author)

  3. Film badge personnel monitoring and dose distribution of industrial and medical workers in the country (1994-1998)

    International Nuclear Information System (INIS)

    Mittal, Hariom; Pandey, R.L.; Massand, O.P.

    2001-03-01

    Personnel Monitoring Section, Bhabha Atomic Research Centre, Mumbai, is entrusted with the responsibility of providing a countrywide personnel monitoring to radiation workers using extemal radiation like X, beta, gamma and neutron. As per Radiation Protection Rules (RPR) of 1971 promulgated by the competent authority the personnel monitoring service is mandatory for all the workers working with radiation. The radiation exposures received by them should be within the limits stipulated by AERB. Nearly 43,000 radiation workers in about 3000 DAE and non DAE institutions using radiation and radioisotopes are monitored, using both the Film Badges and the Thermoluminescent dosimeters. This report presents various aspects of film dosimetry which was introduced more than four decades ago in the country and also analysis of doses received by the radiation workers in the five year block 1994-1998 covered by film badge service. (author)

  4. Neutron measuring instruments for radiation protection

    International Nuclear Information System (INIS)

    Heinzelmann, M.; Schneider, W.; Hoefert, M.; Kuehn, H.; Jahr, R.; Wagner, S.; Piesch, E.

    1979-09-01

    The present report deals with selected topics from the field of neutron dosimetry for radiation protection connected with the work of the subcommittee 6802 in the Standards Committee on Radiology (NAR) of the German Standards Institute (DIN). It is a sort of material collection. The topics are: 1. Measurement of the absorbed-energy dose by a) ionization chambers in fields of mixed radiation and b) recoil-proton proportional counting tubes. 2. Measurement of the equivalent dose, neutron monitors, combination methods by a) rem-meters, b) recoil-proton counting tubes, c) recombination method, tissue-equivalent proportional counters, activation methods for high energies in fields of mixed radiation, d) personnel dosimetry by means of ionization chambers and counting tubes, e) dosimetry by means of activation methods, nuclear track films, nonphotographic nuclear track detectors and solid-state dosimeters. (orig./HP) [de

  5. Adaptation of a free radical dosimeter to the intercomparison of fast neutron beams. Part of a coordinated programme on the development of a transfer instrument for neutron dosimetry intercomparison

    International Nuclear Information System (INIS)

    Bermann, F.

    1978-07-01

    A compressed mixture of alanine and paraffin is a compact and uniform dosemeter with good reproducibility and practical use of it can be made in the range from 10 2 to 10 7 rad. These dosemeters are virtually tissue-equivalent as regards to the absorption of ionizing radiation, especially in the case of fast neutrons. The detection efficiency decreases in the case of particles with high LET. As far as intercomparisons are concerned, this dosemeter shows good relative accuracy of the data and affords an opportunity of successive readings after intervals of time

  6. Development and implementation of a fast neutron monitoring system based on plastic track detectors

    International Nuclear Information System (INIS)

    Bradley, R.P.; Ryan, R.N.

    1988-01-01

    The Bureau of Radiation and Medical devices has provided neutron dosimetry services for Canadian industrial and research applications since the late 1960's. The program used a nuclear emulsion film, Kodak NTA, as the personal dosimeter. Despite its two principle shortcomings, that of a relatively high energy threshold, approximately 700 keV, and highly labor intensive analysis technique, there was little else conveniently available for use. For quite a number of years we pursued, as did many laboratories, the possibilities of developing an albedo dosimeter based on paired thermoluminescent elements in some form of cadmium and plastic encapsulation. Some promise has been shown by this method and several major laboratories have designed and are currently using albedo dosimeters. At the first symposium on Personnel Radiation Dosimetry held in Knoxville in 1984, Dale Hankins of the Lawrence Livermore Laboratories, in his paper entitled, Improvements in the Etching of CR-39 for Large Scale Neutron Dosimetry, reported on his laboratories work with the polycarbonate, CR-39. Using this paper as a start and following up on similar work by W. G. Cross at the Chalk River Nuclear Laboratories, we have developed a replacement dosimeter. This paper will describe the principle features of the system introduced into routine use in Canada in October 1987

  7. Development and underground testing of the α dosimeter: a solid state electronic personal radiation dosimeter for uranium miners

    International Nuclear Information System (INIS)

    Parkinson, R.N.; Roze, V.; Shepherd, R.

    1981-01-01

    The αDOSIMETER is a complete, integrated system designed to monitor the immediate worksite of underground miners where the disintegration for radon daughters is a risk to the health of mining personnel. The dosimeter weighing little more than one pound is worn by each miner throughout the entire shift and is powered by the miner's cap lamp battery. After this integration period, the unit is connected to a reading network whereupon the day's data is dumped, calculated and stored. Beginning in July 1980, prototype units were subjected to vigorous underground testing in uranium mines in Canada and the United States and in tin mines in Cornwall, UK. The testing results are summarized and proposals advanced for a typical mine monitoring system utilizing the αDOSIMETER

  8. Quality assurance manual for the Department of Energy laboratory accreditation program for personnel dosimetry systems

    International Nuclear Information System (INIS)

    1987-02-01

    The overall purpose of this document is to establish a uniform approach to quality assurance. This will ensure that uniform, high-quality personnel dosimetry practices are followed by the participating testing laboratories. The document presents guidelines for calibrating and maintaining measurement and test equipment (M and TE), calibrating radiation fields, and subsequently irradiating and handling personnel dosimeters in laboratories involved in the DOE dosimetry systems testing program. Radiation energies for which the test procedures apply are photons with approximately 15 keV to 2 MeV, beta particles above 0.3 MeV, neutrons with approximately 1 keV to 2 MeV. 12 refs., 4 tabs

  9. A new radiochromic dosimeter film

    Science.gov (United States)

    Sidney, L. N.; Lynch, D. C.; Willet, P. S.

    By employing acid-sensitive leuco dyes in a chlorine-containing polymer matrix, a new radiochromic dosimeter film has been developed for gamma, electron beam, and ultraviolet radiation. These dosimeter films undergo a color change from colorless to royal blue, red fuchsia, or black, depending on dye selection, and have been characterized using a visible spectrophotometer over an absorbed dose range of 1 to 100 kGy. The primary features of the film are improved color stability before and after irradiation, whether stored in the dark or under artificial lights, and improved moisture resistance. The effects of absorbed dose, dose rate, and storage conditions on dosimeter performance are discussed. The dosimeter material may be produced as a free film or coated onto a transparent substrate and optionally backed with adhesive. Potential applications for these materials include gamma sterilization indicator films for food and medical products, electron beam dosimeters, and in-line radiation monitors for electron beam and ultraviolet processing.

  10. Thermoluminescence and phosphate glass dosimeter systems in the low dose range

    International Nuclear Information System (INIS)

    Piesch, E.; Burgkhardt, B.

    1978-06-01

    This report describes a standard test program for TLD and RPL systems worked out by the Working Party on 'Dose Measurement of External Radiation' by the Fachverband fuer Strahlenschutz e.V. to demonstrate the performance of dosimeter systems to be employed in environmental monitoring and in personnel dosimetry. The results of an intercomparison study are outlined in which 17 laboratories from the German speaking countries participated with 43 dosimeter systems. (orig.) [de

  11. Procedure for the delivering of personal short-term visitor dosimeters

    CERN Document Server

    2016-01-01

    Update of the administrative procedure for delivering a personal short-term visitor dosimeter to associated members of CERN’s personnel.   Associated members of the CERN personnel may request a short term visitor dosimeter if working only in Supervised Radiation Areas and for a period of less than two months in a calendar year. Such a dosimeter is delivered without the need to provide the usual regular documents: radiation passport, certificate from the home institute or medical certificate. Periodic verification will ensure that holders of these personal dosimeters do not exceed the maximum allowed personal dose for this type of dosimeter, which is the same as the limit for members of the public at 1 mSv per year. From now on, the two-month period can be spread over a calendar year, offering greater flexibility to users coming to CERN for multiple short periods. Please return unused dosimeters Persons leaving CERN for a period of more than one month should return their dosimeter to the D...

  12. Direct reading dosimeter

    International Nuclear Information System (INIS)

    Thomson, I.

    1985-01-01

    This invention is a direct reading dosimeter which is light, small enough to be worn on a person, and measures both dose rates and total dose. It is based on a semiconductor sensor. The gate threshold voltage change rather than absolute value is measured and displayed as a direct reading of the dose rate. This is effected by continuously switching the gate of an MOS transistor from positive to negative bias. The output can directly drive a digital readout or trigger an audible alarm. The sensor device can be a MOSFET, bipolar transistor, or MOSFET capacitor which has its electrical characteristics change due to the trapped charge in the insulating layer of the device

  13. Colorimetric gas dosimeter

    International Nuclear Information System (INIS)

    McConnaughey, P.W.; McKee, E.S.

    1984-01-01

    A gas dosimeter comprises a stack of porous sheets, impregnated with a reagent that changes color on contact with the gas to be determined, contained in a housing which has an opening to expose one end of the stack to the atmosphere to be tested. The gas to be determined penetrates by diffusion the layers of porous sheets, causing the sheets in the stack to change color sequentially from the end of the stack exposed to the atmosphere. The degree of penetration through the layers of porous sheets is a function of dosage exposure. The housing may be transparent with each superposed sheet in the stack being larger than the adjacent underlying sheet, so that each sheet is visible through the housing endwall

  14. The intelligence of dosimeter for ionization radiation

    International Nuclear Information System (INIS)

    He Jinglun

    1992-01-01

    The connection of dosimeter with microcomputer system is described, which has the functions of sampling, data handling, display and printing dose values in legal units of measurement. The accuracy and speed of measurement for dosimeters are also raised, thereby the dosimeters are made to have intelligence and the application range of dosimeter is enlarged

  15. Quo vadis, personnel monitoring

    International Nuclear Information System (INIS)

    Becker, K.

    1975-01-01

    With the increasing use of nuclear power and radiation sources, the selection of optimum systems for personnel monitoring is becoming a matter of worldwide concern. The present status of personnel dosimetry, sometimes characterized by unstable and inaccurate detectors and oversimplified interpretation of the results, leaves much to be desired. In particular, photographic film, although having certain advantages with regard to economics and information content, undergoes rapid changes in warm and humid climates. Careful sealing reduces, but does not prevent, these problems. The replacement of film by solid-state dosimeters, primarily thermoluminescence dosimeters, is in progress or being considered by an increasing number of institutions and requires a number of decisions concerning the choice of the optimum detector(s), badge design, and evaluation system; organizational matters, such as the desirability of automation and computerized bookkeeping; etc. The change also implies the potential use of such advanced concepts as different detectors and monitoring periods for the large number of low-risk persons and the small number of high-risk radiation workers. (auth)

  16. Silver nitrate based gel dosimeter

    International Nuclear Information System (INIS)

    Titus, D; Samuel, E J J; Srinivasan, K; Roopan, S M; Madhu, C S

    2017-01-01

    A new radiochromic gel dosimeter based on silver nitrate and a normoxic gel dosimeter was investigated using UV-Visible spectrophotometry in the clinical dose range. Gamma radiation induced the synthesis of silver nanoparticles in the gel and is confirmed from the UV-Visible spectrum which shows an absorbance peak at around 450 nm. The dose response function of the dosimeter is found to be linear upto12Gy. In addition, the gel samples were found to be stable which were kept under refrigeration. (paper)

  17. Evaluation of personal integrating dosimeters

    International Nuclear Information System (INIS)

    Correa, C.A.; Bisauta, Mauricio A.

    2007-01-01

    The objective of this work is to analyze the different types of dosimeters present in the international market that are used to provide personal dose monitoring, specifically for external gamma and beta radiation, Hp(10) and Hp (0,07), as well as to add comments of advances in the field of passive and operative dosimetry, and the changes that are being produced in the regulating policy in other countries regarding the use of this devices. The technical specification of each dosimeter has been extracted of different catalogues of products. To conclude, the importance has been stressed in a proper selection of dosimeters with its advantages and disadvantages before its use. (author) [es

  18. Reactor neutron dosimetry

    International Nuclear Information System (INIS)

    Najzer, M.; Pauko, M.; Glumac, B.; Acquah, I.N.; Moskon, F.

    1977-01-01

    An analysis of requirements and possibilities for experimental neutron spectrum determination during the reactor pressure vessel surveil lance programme is given. Fast neutron spectrum and neutron dose rate were measured in the Fast neutron irradiation facility of our TRIGA reactor. It was shown that the facility can be used for calibration of neutron dosimeters and for irradiation of samples sensitive to neutron radiation. The investigation of the unfolding algorithm ITER was continued. Based on this investigations are two specialized unfolding program packages ITERAD and ITERGS written this year. They are able to unfold data from activation detectors and NaI(T1) gamma spectrometer respectively

  19. Heater design for reading radiation dosimeters

    International Nuclear Information System (INIS)

    Seidel, J.G.; Felice, P.E.

    1982-01-01

    The nichrome heating element of a conventional dosimeter reading apparatus has been redesigned to include a flat-bottomed depression big enough to hold a thermoluminescent dosimeter. A thin glass plate is positioned in the recess on top of the dosimeter to retain it in the recess during the heating and reading process. This technique of securing the dosimeter in contact with the heating element avoids physical scratching or damage to the dosimeter

  20. Citizen's dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Klemic, Gladys [Naperville, IL; Bailey, Paul [Chicago, IL; Breheny, Cecilia [Yonkers, NY

    2008-09-02

    The present invention relates to a citizen's dosimeter. More specifically, the invention relates to a small, portable, personal dosimetry device designed to be used in the wake of a event involving a Radiological Dispersal Device (RDD), Improvised Nuclear Device (IND), or other event resulting in the contamination of large area with radioactive material or where on site personal dosimetry is required. The card sized dosimeter generally comprises: a lower card layer, the lower card body having an inner and outer side; a upper card layer, the layer card having an inner and outer side; an optically stimulated luminescent material (OSLM), wherein the OSLM is sandwiched between the inner side of the lower card layer and the inner side of the upper card layer during dosimeter radiation recording, a shutter means for exposing at least one side of the OSLM for dosimeter readout; and an energy compensation filter attached to the outer sides of the lower and upper card layers.

  1. Sensitive chemical neutron dosimetry using silver colloids

    International Nuclear Information System (INIS)

    Brede, O.; Boes, J.; Hoesselbarth, B.

    1982-01-01

    The radiation-induced formation of silver colloid was checked for its use as a sensitive dosimeter for neutron irradiation. For non-monoenergetic pulsed neutron irradiation in the Dubna IBR-30 reactor, the colloid dosimeter was found to be suitable to indicate the chemical neutron effect, i.e., to determine the sum concentration of the primary particles of water radiolysis: esub(aq)sup(-), OH and H. (author)

  2. Review of Fricke gel dosimeters

    International Nuclear Information System (INIS)

    Schreiner, L J

    2004-01-01

    The innovation of adding a gel matrix to the traditional Fricke dosimeter to stabilize geometric information established the field of gel dosimetry for radiation therapy. A discussion of Fricke gels provides an overview of the issues that determine the dose response of all gel dosimeters in general. In this paper we review some of the features of Fricke systems to illustrate these issues and, in addition, to motivate renewed clinical interest in Fricke gels

  3. Method to improve the evaluation of a combination track-etch dosimeter/spectrometer

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Parkhurst, M.A.; Hadlock, D.E.; Faust, L.G.

    1983-09-01

    A paper is summarized which describes a method of determining the neutron energy spectrum through spectrum unfolding techniques to more accurately assess the dose equivalent from track-etch dosimeters. A mathematical technique is described which can be used in conjunction with the neutron detectors to more accurately estimate neutron dose equivalent. The technique is based upon solutions to a system of Fredholm integral equations of the first type

  4. The Calvet calorimetric dosimeter

    International Nuclear Information System (INIS)

    Puig, J.R.; Romano, F.

    1965-01-01

    This report describes a dosimeter based on the conduction calorimetry principle, and designed to operate in swimming-pool type nuclear reactors. The properties of the apparatus are as follows: 1 - the measurement is independent of the specific heat of the calorimetric elements; 2 - each calorimetric element is fitted with an electrical calibration; 3 - the apparatus is made up of two independent calorimetric elements; 4 - the nature of the calorimetric elements makes it possible to analyse the radiation received; 5 - the measurable intensities of the absorbed radiation vary from 4 to 4000 M/rads per hour; 6 - the sensitive part of the apparatus is fitted inside a cylinder 5 cm high and 2 cm in diameter. One pre-production unit made up of graphite and beryllium cores has been tried out in the reactor Siloe with radiation intensities of about 1 to 2 watts per gram. It absorbed an accumulated dose of 1.2*1O 12 rads without any weaknesses appearing. (authors) [fr

  5. Study on the angular dependence of personal exposure dosimeter - Focus on thermoluminescent dosimeter and photoluminescent dosimeter

    International Nuclear Information System (INIS)

    Dong, Kyung-Rae; Kweon, Dae Cheol; Chung, Woon-Kwan; Goo, Eun-Hoe; Dieter, Kevin; Choe, Chong-Hwan

    2011-01-01

    Radiation management departments place more emphasis on the accuracy of measurements than on the increase in the average dose and personal exposure dose from the use of radiation equipment and radioactive isotopes. Although current measurements are taken using devices, such as film badge dosimeters, pocket dosimeters and thermoluminescent dosimeters (TLDs), this study compared the angular dependence between the widely used TLDs and photoluminescent dosimeter (PLDs) in order to present primary data and evaluate the utility of PLD as a new dosimeter device. For X-ray fluoroscopy, a whole body phantom was placed on a table with a setting for the G-I technical factors fixed at a range of approximately 40 cm with a range of ±90 o at an interval scale of 15 o from the center location of an average radiological worker for PLDs (GD-450) and TLDs (Carot). This process was repeated 10 times, and at each time, the cumulative dosage was interpreted from 130 dosimeters using TLDs (UD-710R, Panasonic) and PLDs (FGD-650). The TLD and PLD showed a 52% and 23% decrease in the depth dosage from 0 o to -90 o , respectively. Therefore, PLDs have a lower angular dependence than TLDs.

  6. Personnel radiation monitoring by thermoluminescence dosimetry (2000-2001)

    International Nuclear Information System (INIS)

    Mi-Cho-Cho, Daw; Hla-Hla-Win, Daw; Thin-Thin-Kraing

    2001-01-01

    Thermoluminescence dosimetry service was introduced in 1991. Personnel who exposed directly or indirectly to radiation is monitored by thermoluminescent dosimeter. TL materials used for thermoluminescent dosimeter are in the form of thin disc. Personnel whole-body and extremity doses are measured by thermoluminescence dosimetry. The Harshaw Model 4500 TLD reader and Vinten 654E TLD reader are used for TLD evaluation. At present about 600 radiation workers are provided with TLD for routine monitoring. It was found that most personnel had received within permissible dose recommended by the International Commission on Radiological Protection (ICRP). (author)

  7. Design and development of a PMOSFET gamma ray dosimeter

    International Nuclear Information System (INIS)

    Khanna, V.K.; Kumar, A.; Gupta, R.P.; Pandya, A.; Roy, Rajesh

    2005-01-01

    A P-channel MOSFET chip has been designed for detection of gamma radiations. The chip consists of three MOSFETs of different geometrical parameters for achieving sensitivity to low and high dose ranges. One of the MOSFET structures has a closed geometry to reduce the leakage current. The developed dosimeter being a MOSFET, its IC (Integrated Circuit)-compatibility helps in easy interfacing with readout circuitry. The dosimeter fabrication process is based on metal-gate MOSFET technology with thick gate oxide to increase the effective number of electron-hole pairs generated by the gamma rays impinging on the device. The process for the chip realization has been designed and simulated to achieve the required impurity diffusion profile. The chip has been fabricated using the above process and electrically characterized. The device has been exposed to gamma ray source and its characteristics measured. The change in threshold voltage of the MOSFET after exposure has been used to calculate the sensitivity of the device. The developed dosimeter has potential applications in personnel dosimetry and cancer treatment. This paper describes the basic detection mechanism of the MOSFET, the design approach, and fabrication process of the MOSFET dosimeter. (author)

  8. Personal dosimetry in a mixed field of high energy muons and neutrons

    International Nuclear Information System (INIS)

    Cossairt, J.D.; Elwyn, A.J.

    1986-11-01

    High energy accelerators quite often emit muons. These particles behave in matter as would heavy electrons and are thus difficult to attenuate with shielding in many situations. Hence, these muons can be a source of radiation exposure to personnel and suitable methods of measuring the absorbed dose received to these people is obviously required. In practical situations, such muon radiation fields are often mixed with neutrons, well-known to be an even more troublesome particle species with respect to dosimetry. In this paper, we report on fluence measurements made in such a mixed radiation field and a comparison of dosimeter responses. We conclude that commercial self-reading dosimeters and film badges provided an adequate measure of the absorbed dose due to muons

  9. Electronic dosimeter characteristics and new developments

    International Nuclear Information System (INIS)

    Thompson, I.M.G.

    1999-01-01

    Electronic dosimeters are very much more versatile than existing passive dosimeters such as TLDs and film badges which have previously been the only type of dosimeters approved by national authorities for the legal measurement of doses to occupationally exposed workers. Requirements for the specifications and testing of electronic dosimeters are given in the standards produced by the International Electrotechnical Commission Working Group IEC SC45B/B8. A description is given of these standards and the use of electronic dosimeters as legal dosimeters is discussed. (author)

  10. Mexican gems as thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Azorin N, J.

    1979-01-01

    The possibility of using naturally ocurring mexican gems as thermoluminescent dosimeters (TLD) was investigated. Twelve types of gems were irradiated with X and gamma rays in order to determinate their dosimetric properties. Three of these gems showed favorable thermoluminescent characteristics compared with commercial thermoluminescent dosimeters. The plots of their thermoluminescent response as a function of gamma dose are straight lines on full log paper in the dose range 10 -2 to 10 2 Gy. The energy dependence is very strong to low energies of the radiation. Their fading was found to be about 5%/yr. and they may be annealed as reused without loss in sensitivity. Therefore, these gems can be used as X and gamma radiation dosimeters. (author)

  11. CVD diamond detectors and dosimeters

    International Nuclear Information System (INIS)

    Manfredotti, C.; Fizzotti, F.; LoGiudice, A.; Paolini, C.; Oliviero, P.; Vittone, E.; Torino Univ., Torino

    2002-01-01

    Natural diamond, because of its well-known properties of tissue-equivalence, has recorded a wide spreading use in radiotherapy planning with electron linear accelerators. Artificial diamond dosimeters, as obtained by Chemical Vapour Deposition (CVD) could be capable to offer the same performances and they can be prepared in different volumes and shapes. The dosimeter sensitivity per unit volume may be easily proved to be better than standard ionization microchamber. We have prepared in our laboratory CVD diamond microchamber (diamond tips) in emispherical shape with an external diameter of 200 μm, which can be used both as X-ray beam profilometers and as microdosimeters for small field applications like stereotaxy and also for in vivo applications. These dosimeters, which are obtained on a wire substrate that could be either metallic or SiC or even graphite, display good performances also as ion or synchrotron X-rays detectors

  12. To the attention of all dosimeter users

    CERN Multimedia

    2005-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include a compulsory monthly read-out of the dosimeter. Therefore we would like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350.-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short term visitor dosimeter (VCT). This dosimeter has a limited validity period but without for a compulsory periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service Bldg. 24 E 011 http://cern.ch/rp-dosimetry

  13. To the attention of all dosimeter users

    CERN Multimedia

    Dosimetry Service

    2005-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include the compulsory monthly read-out of the dosimeter. Therefore we would like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350.-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short term visitor dosimeter (VCT). This dosimeter has a limited validity period but without for a compulsory periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service Bld 24 E 011 http://cern.ch/rp-dosimetry

  14. To the attention of all dosimeter users

    CERN Multimedia

    Dosimetry Service

    2006-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include the compulsory monthly read-out of the dosimeter. We would therefore like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350,-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short-term visitor dosimeter (VCT). This dosimeter has a limited validity period but does not require a periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service - Bldg. 24 E 011 - http://cern.ch/rp-dosimetry

  15. Flame-sintered ceramic exoelectron dosimeter samples

    International Nuclear Information System (INIS)

    Petel, M.; Holzapfel, G.

    1979-01-01

    New techniques for the preparation of integrating solid state dosimeters, particularly exoelectron dosimeters, have been initiated. The procedure consists in melting the powdered dosimeter materials in a hot, fast gas stream and depositing the ceramic layer. The gas stream is generated either through a chemical flame or by an electrical arc plasma. Results will be reported on the system Al 2 O 3 /stainless steel as a first step to a usable exoelectron dosimeter

  16. Dosimeter charging and/or reading apparatus

    International Nuclear Information System (INIS)

    Fine, L.T.; Jackson, T.P.

    1980-01-01

    A device is disclosed for charging and/or reading a capacitor associated with an electrometer incorporated in a radiation dosimeter for the purpose of initializing or ''zeroing'', the dosimeter at the commencement of a radiation measurement cycle or reading it at any time thereafter. The dosimeter electrometer has a movable electrode the position of which is indicative of the charge remaining on the dosimeter capacitor and in turn the amount of radiation incident on the dosimeter since it was zeroed. The charging device also includes means for discharging, immediately upon conclusion of the dosimeter capacitor charging operation, stray capacitance inherent in the dosimeter by reason of its mechanical construction. The charge on the stray capacitance, if not discharged at the conclusion of the dosimeter capacitor charging operation, leaks off during the measurement cycle, introducing measurement errors. A light source and suitable switch means are provided for automatically illuminating the movable electrode of the dosimeter electrometer as an incident to charging the dosimeter capacitor to facilitate reading the initial, or ''zero'', position of the movable electrometer electrode after the dosimeter capacitor has been charged and the stray capacitance discharged. Also included is a manually actuatable switch means, which is operable independently of the aforementioned automatic switch means, to energize the lamp and facilitate reading of the dosimeter without charging

  17. Personnel training and certification

    International Nuclear Information System (INIS)

    Whittemore, W.L.

    1976-01-01

    In order to make the full benefits of neutron radiography available in the nondestructive test (NDT) field, it has been necessary to formalize its application. A group under the Penetrating Radiation Committee of the American Society for Nondestructive Testing (ASNT) was organized to prepare a recommended practice for neutron radiography. The recommended practices require the establishment of an appropriate certification program. The requirements on the employer to establish and maintain a qualification and certification program are outlined. To conduct a program of nondestructive testing using neutron radiography requires the usual three levels of qualified and certified personnel. The program is administered by a Level III person. Routine exposure, reviews, and reporting of test results are the responsibilities of Level I and Level II personnal. The amount of training and nature of the required practical examination are also specified. The recommended practices document assures users that NDT work in the field of neutron radiography is performed by qualified personnel. Although no training courses are available to provide experience in the depth required by the recommended practices document, SNT-TC-1A, short courses are provided at a number of locations to familarize user's representatives with the interpretation of neutron radiographs and capabilities and limitations of the technique

  18. Area monitoring dosimeter program for the Pacific Northwest National Laboratory: Results for CY 1997

    International Nuclear Information System (INIS)

    Bivins, S.R.; Stoetzel, G.A.

    1998-07-01

    In January 1993, Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the US Department of Energy (DOE) Radiological Control Manual (RCM). The purpose of the program was to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a) (1)--(3) and Article 511.1 of the RCM, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually, and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years (CY) 1993--1996 confirmed that personnel dosimetry was not needed for individuals located in areas monitored by the program. A total of 93 area thermoluminescent dosimeters (TLDs) were placed in PNNL facilities during CY 1997. The TLDs were exchanged and analyzed quarterly. All routine area monitoring TLD results were less than 50 mrem annually after correcting for worker occupancy. The results support the conclusions that personnel dosimeters are not necessary for staff, declared pregnant workers, minors, or members of the public in these monitored areas

  19. Area Monitoring Dosimeter Program for the Pacific Northwest National Laboratory: Results for CY 1998

    International Nuclear Information System (INIS)

    GA Stoetzel; SR Bivins

    1999-01-01

    In January 1993, Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the Department of Energy (DOE) Radiological Control Manual (RCM). The purpose of the program was to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a)(1)-(4) and Article 511.1 of the RCM, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years 1993-1997 confirmed that personnel dosimetry was not needed for individuals located in areas monitored by the program. A total of 97 area thermoluminescent dosimeters (TLDs) were placed in PNNL facilities during calendar year 1998. The TLDs were exchanged and analyzed quarterly. All routine area monitoring TLD results were less than 50 mrem annually after correcting for worker occupancy. The results support the conclusion that personnel dosimeters are not necessary for staff, declared pregnant workers, minors, or members of the public in these monitored areas

  20. Do you know DIS? a novel passive individual dosimeterd. Direct Ion Storage dosimeter DIS-1 officially approved in Switzerland

    International Nuclear Information System (INIS)

    Fiechtner, A.; Wernli, C.

    2001-01-01

    For individual monitoring film and TLD are the most often used types of dosimeters. On a smaller scale phosphate glasses and detectors based on optically stimulated luminescence (OSL) are also in use. As a new addition to the list of available personnel dosimeters the direct ion storage (DIS) system became legally approved for the first time in Switzerland. At the Paul Scherrer Institute (PSI) the RADOS dosimetry system DIS-1 is in official use since beginning of 2001. (orig.) [de

  1. Neutron detection and characterization for non-proliferation applications using 3D computer optical memories [Use of 3D optical computer memory for radiation detectors/dosimeters. Final progress report

    International Nuclear Information System (INIS)

    Phillips, Gary W.

    2000-01-01

    We have investigated 3-dimensional optical random access memory (3D-ORAM) materials for detection and characterization of charged particles of neutrons by detecting tracks left by the recoil charged particles produced by the neutrons. We have characterized the response of these materials to protons, alpha particles and carbon-12 nuclei as a functions of dose and energy. We have observed individual tracks using scanning electron microscopy and atomic force microscopy. We are investigating the use of neural net analysis to characterize energetic neutron fields from their track structure in these materials

  2. Improved sample holders for the PMMA dosimeters

    International Nuclear Information System (INIS)

    Kobayashi, Toshikazu; Sone, Koji; Iso, Katsuaki

    1994-01-01

    PMMA dosimeters are widely used for high dose dosimetry. Dose is determined by measuring the change in optical density of the irradiated PMMA dosimeter element. Measurement precision depends on the mounting method of a dosimeter element in the sample room of a spectrophotometer. We tried to prepare three types of holders, (holders A, B and C in Figs. 1-3), according to the shape of PMMA dosimeter elements. We measured optical density of the irradiated PMMA dosimeter elements by using the three types of holders. It is revealed that the holder of the type A gives more precise results for the Red 4034 or Gammachrome YR dosimeter than that of the type B. The measurements with a spectrophotometer using the type C holder gives better results for the Red acrylic dosimeter than the case of the measurements by the exclusive reader. (author)

  3. Neutron dosimetry in containment of a pressurized water reactor utilizing the Panasonic UD-802 dosimetry system

    International Nuclear Information System (INIS)

    Kralick, S.C.

    1984-01-01

    The Panasonic UD-802 dosimeter was evaluated as a potential neutron dosimeter for use in containment of a PWR. The Panasonic UD-802 dosimeter, although designed as a beta and gamma dosimeter, is also sensitive to neutrons. UD-802 dosimeters were mounted on polyethylene phantoms and irradiated to known doses at selected locations in containment. The known neutron dose equivalents were determined based on remmeter dose rate measurements and stay times. The thermoluminescent response of the dosimeters and the known neutron dose equivalents were used to obtain a calibration factor at each location. The average calibration factor was 3.7 (unit of dosimeter response per mrem) and all calibration factors were within +-30% of this mean value. The dosimeter distance from the phantom was found to have minimal effect on the response but the system was directionally dependent, necessitating a correction in the calibration factor. The minimum significant dosimeter response was determined independent of any calibration factor. The minimum significant response of the UD-802 to neutrons is a function of the corresponding gamma exposure rate. It is concluded that the Panasonic UD-802 dosimeter can be used for neutron dosimetry in PWR containment

  4. Suitability of CR-39 dosimeters for personal dosimetry around CANDU reactors

    International Nuclear Information System (INIS)

    Cross, W.G.

    1992-08-01

    The capabilities and limitations of CR-39 damage track detectors have been evaluated for their use as personal neutron dosimeters around CANDU reactors. Since the energy response is a critical characteristic, the neutron energy spectra expected within CANDU containments were studied. In the boiler rooms, around the moderator cooling systems, and in most of the fueling machine vaults, the spectra vary considerably, but the majority of the dose is expected to be delivered by neutrons above 80 keV, the approximate threshold for electrochemically-etched CR-39 detectors. In the Pickering A fueling machine vault, and in areas in other stations to which neutrons from reactors have been multiply scattered, lower energy neutrons may be important. In nearly all areas where people work, it appears that working times will be limited by gamma rays rather than by neutrons. The characteristics of other neutron dosimeters - bubble and superheated drop detectors, albedo detectors, and Si real-time detectors - were also reviewed. For workers who typically receive neutron doses that are small compared with regulatory limits, CR-39 is the most suitable available dosimeter for demonstrating compliance. All single dosimeters have poor angular response over the range 0 to 180 degrees because of the shielding of the body. Albedo and Si detectors have particularly poor energy responses over the energy range of importance. Bubble and superheated drop detectors have the advantages of immediate readout and high sensitivity, but the disadvantages of inability to integrate doses over a long period, temperature dependence, very limited range and higher cost. (Author) (110 refs., 45 figs.)

  5. Bronchial dosimeter for radon progeny

    Energy Technology Data Exchange (ETDEWEB)

    Cheung, T.K.; Yu, K.N.; Nikezic, D.; Haque, A.K.M.M. [City University of Hong Kong, Hong Kong (China); Vucic, D. [Faculty of Technology, University of Nis, Lescovac (Yugoslavia)

    2000-05-01

    Traditionally, assessments of the bronchial dose from radon progeny were carried out by measuring the unattached fraction (f{sub p}) of potential alpha energy concentration (PAEC), the total PAEC, activity median diameters (AMDs) and equilibrium factor, and then using dosimetric lung models. A breakthrough was proposed by Hopke et al. (1990) to use multiple metal wire screens to mimic the deposition properties of radon progeny in the nasal (N) and tracheobronchial (T-B) regions directly. In particular, they were successful in using four layers of 400-mesh wire screens with a face velocity of 12 cm s{sup -1} for the simulation of radon progeny deposition in the T-B region. Oberstedt and Vanmarcke (1995) carried out precise calibrations for the system, and named the system as the 'bronchial dosimeter'. Based on these, Yu and Guan (1998) proposed a portable bronchial dosimeter similar to a normal measurement system for radon progeny or PAEC and consisted of only a single sampler and employed only one 400-mesh wire screen and one filter. However, all these 'bronchial dosimeters' in fact only determined the fraction of potential alpha energy from radon progeny deposited in the T-B region, which required certain assumptions and calculations to further give the final bronchial dose. In the present work, a true 'bronchial dosimeter' was designed, which consisted of three 400-mesh wire screens and a filter. With a face velocity of 11 cm s{sup -1}, the deposition pattern on the wire screens was found to satisfactorily match the variation of the dose conversion factor (in the unit of mSv/WLM) with the size of radon progeny from 1 to 1000 nm. In this way, this bronchial dosimeter directly gave the bronchial dose from the alpha counts recorded on the wire-screens and the filter paper. With the development of this bronchial dosimeter, the present practice of 'dose estimation' from large-scale radon surveys can be replaced by large

  6. Operation of Personal Electronic Dosimeters at NRCN

    International Nuclear Information System (INIS)

    Weinstein, M.; Abraham, A.; Tshuva, A.; German, U.

    2004-01-01

    In the recent years, electronic personal dosimeters (EPD's) are increasingly being used at NRCN, replacing the old direct reading dosimeters that are still widely used. The most significant advantage of the new dosimeters is the real time alarm in a radiation field exceeding a pre-determined threshold. The EPD dosimeters are more precise and can measure γ, β and x rays of a wide range of energies. In addition, the electronic dosimeters collects and stores the reading at a fixed pattern (every 10 seconds) and keeps the data until it is downloaded from the dosimeter. This feature gives the ability to build a personal time-dependent exposure report for each worker who carries this device and to analyze, identify and measure the exact dose, time and duration of any exposure event he was involved in. Designing and building a personal electronic dosimeter became possible as a result of the massive technological improvements of semi conductor detectors and the minimization processes of microprocessors and low energy electronic devices. The main purpose for personal electronic dosimeters was to monitor on-line doses for radiation workers.A special reader device enables to download data and upload operational settings of the dosimeters. By means of this communication channel, one can save the data acquired by the dosimeter, clear the dosimeter memory and set the dosimeter operational parameters. There are two possible working patterns. The first is to read and set all the dosimeters at a central point, normally a dosimetry laboratory (single reader) and the second and more expensive one, is to build a network of readers covering the plant for obtaining on-line communication

  7. Development of colored alumilite dosimeter

    CERN Document Server

    Obara, K; Yagi, T; Yokoo, N

    2003-01-01

    In the ITER (International Thermonuclear Experimental Reactor), in-vessel components such as blanket and divertor, which are installed in the vacuum vessel of the ITER, are maintained by remote handling equipment (RH equipment). The RH equipment for maintenance is operated under sever environmental conditions, such as high temperature (50 approx 100 degC), high gamma-ray radiation (approx 1 kGy/h) in an atmosphere of inert gas or vacuum; therefore many components of the RH equipment must have a suitable radiation resistance efficiency for long time operation (10 approx 100 MGy). Typical components of the RH equipment have been extensively tested under an intensive gamma-ray radiation. Monitoring of the radiation dose of the components of the RH equipment is essential to control the operation period of the RH equipment considering radiation resistance. However, the maximum measurable radiation dose of the conventional dosimeters, such as ionization chamber, liquid, glass and plastic dosimeters are limited to b...

  8. Personnel monitoring in geologic fields

    International Nuclear Information System (INIS)

    Romanova, I.N.; Seredin, Yu.V.

    1981-01-01

    State of radiation safety for the personnel of geologic crews carrying out neutron logging of wells using Po-Be sources has been evaluated. Given are results of development of methods for the evaluation of individual radiation loads for personnel when working with Po-Be neutron sources useful for the application in practice by a geologic logging crew as well as a quantitative evaluation of profissional radiation loads during this kind of work. The following methods are recommended for personnel monitoring: 1) calculation of whole-body irradiation doses and hands from averaged values of radiation dose rate; 2) calculational tabulated determination of irradiation doses during recharging of shanks of well instruments. Personnel monitoring by means of instrumental methods is not necessary in the considered case [ru

  9. RADIATION DOSIMETER AND DOSIMETRIC METHODS

    Science.gov (United States)

    Taplin, G.V.

    1958-10-28

    The determination of ionizing radiation by means of single fluid phase chemical dosimeters of the colorimetric type is presented. A single fluid composition is used consisting of a chlorinated hydrocarbon, an acidimetric dye, a normalizer and water. Suitable chlorinated hydrocarbons are carbon tetrachloride, chloroform, trichloroethylene, trichlorethane, ethylene dichioride and tetracbloroethylene. Suitable acidimetric indicator dyes are phenol red, bromcresol purple, and creosol red. Suitable normallzers are resorcinol, geraniol, meta cresol, alpha -tocopberol, and alpha -naphthol.

  10. Approving of personal dosimeter services

    International Nuclear Information System (INIS)

    Bergman, K.; Malmqvist, L.

    2001-09-01

    The Swedish regulation SSI FS 98:5 requires that radiological workers of category A use dosemeters from an approved personal dosimetry service. The regulation also includes certain specific dosimeter requirements, which are based on those presented in the Technical Recommendations by the European Commission (Report EUR 14852 EN, 1994). All services have been tested for their ability to determine Hp(10) and some of them to determine Hp(0.07) at one radiation quality. The test was performed in the interval 0.2 mSv to 100 mSv at three different dose equivalents unknown to the system owner. The 11 services operating in Sweden at the moment use 5 different types of dosimeters. The five unique systems have been tested regarding the angular and energy dependence of the response of the dosimeters. The dosimeters were irradiated to a personal dose equivalent of about 1 mSv at three photon energies and at four angles (0, 20, 40 and 60 deg. resp. ) both vertically and horizontally rotated. Only 2 of the services determine Hp(0.07) for beta and gamma radiation and were tested for this quantity. The test results for Hp(10) are all except two within the trumpet curve. For the unique systems it is shown that the uncertainty related to angular response at a specified energy is within the required ±40 % except for the lowest X-ray quality at 40 kV. The response is more dependent on photon energy than on the direction of the photon radiation and the choice of radiation quality for the calibration is of great importance for the system performance

  11. Radiation sensitive polymer gel dosimeters

    International Nuclear Information System (INIS)

    Lepage, M.; Back, S.A.J.; Baldock, C.; Whittaker, A.K.; Rintoul, L.

    2000-01-01

    Full text: Radiation sensitive gels are studied for their potential to retain a permanent 3D dose distribution for applications in radiotherapy. Co-monomers dissolved in a tissue-equivalent hydrogel undergo a polymerization reaction upon absorption of ionizing radiation. The polymer formed influences the local spin-spin relaxation time (T 2 ) of the dosimeter that can be determined using magnetic resonance imaging (MRI). The relationship between T2 and the absorbed dose was studied for different initial chemical compositions. The aim was to find a model linking the changes in T 2 with absorbed dose to the initial composition of the dosimeter. It is believed this will help designing new gel dosimeters having desired properties to minimize the uncertainty in the determination of the dose distribution. 1 H, 13 C nuclear magnetic resonance spectroscopy and FT-Raman spectroscopy were used to quantify the amount of monomers still remaining after the absorption of a given dose of radiation. This data is used to model the changes of T2 as a function of the absorbed dose. A model of fast exchange of magnetization between three proton pools was used, where the fraction of protons (f x H ) in the x th pool is obtained from the chemical composition of the dosimeter and the apparent T2 of each pool is determined for a given composition. Initially, the protons are contained in two pools; a mobile (mob), which contains the water protons and the monomers protons, and a gelatin (gela) proton pool. The mobile pool is partially depleted as polymer is formed, the protons are transferred into the polymer (pol) pool. In the figure, the experimental data along with the calculated values are plotted for three different monomer concentrations, with the gelatin concentration fixed. The model is seen to provide a good fit to the experimental data

  12. Area Monitoring Dosimeter Program for the Pacific Northwest National Laboratory: Results for CY 2000

    International Nuclear Information System (INIS)

    Bivins, Steven R; Stoetzel, Gregory A

    2001-01-01

    Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the Department of Energy (DOE) Radiological Control Manual (RCM) in January 1993. This program is to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a) (1)-(4) and Article 511.1 of the DOE Standard Radiological Control, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually, and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years 1993-2000 confirm that personnel dosimetry is not needed for individuals located in areas monitored by the program

  13. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR

    International Nuclear Information System (INIS)

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez

    2013-01-01

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba

  14. measurement of the supralinearity of 7LiF thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Hancock, I.B.

    1978-04-01

    A study has been made of the supralinearity of the 7 LiF thermoluminesecent dosimeter (T.L.D.) used in the UKAEA personnel criticality dosimeter by exposure to a 60 Co source. the response of the T.L.D. was linear up to approximately 450 rad, above which it became increasingly supralinear, over-reading by about 50% at 2000 rad. The results have been used to produce a correction graph, as a function of apparent dose up to 3500 rad, for routine use in nuclear accident dosimetry. (author)

  15. The status of ANSI N13.11 - The dosimeter performance test standard

    International Nuclear Information System (INIS)

    Sims, C.S.

    1991-01-01

    The standard designated ANSI N13.11 was issued in 1983 and is the basis for the dosimeter performance test associated with the National Voluntary Laboratory Accreditation Program (NVLAP). The standard is important because the Nuclear Regulatory Commission requires that all licensees use personnel dosimeters processed by NVLAP-accredited processors. The standard has undergone review and modifications have been recommended. Historical information concerning the development and utilization of the present standard, ANSI N13.11-1983, is presented. Details associated with the review of the standard (e.g., policy, group selection) are then given. Finally, the modifications recommended by the review group are discussed

  16. Argentine Republic intercomparison programme for personal dosimeters

    International Nuclear Information System (INIS)

    Gregori, Beatriz N.; Papadopulos, Susana B.; Kunst, Juan J.; Cruzate, Juan A.; Saravi, Margarita C.

    2004-01-01

    In 1997 an Intercomparison Program for individual monitoring started in order to test (on a voluntary basis) the performance in absorbed dose and personal dose equivalent determinations. The aim of the program was also to gain some insight into the general aspects related to the type of the personnel dosimeter used, the calibration procedures, the phantom spectral dependence and the management of radiological quantities. The Regional Reference Center for Dosimetry (CCR), of the Argentine National Atomic Energy Commission and the Physical Dosimetry Laboratory of the Argentine National Regulatory Authority, performed the irradiations. Those were done free air and on ICRU phantom, using X-ray, quality ISO: W60, W80, W110 and W200; and 137 Cs and 60 Co gamma rays, normal and angular (0, 30, 60 degrees) incidence. In the framework of the Program, an upgraded workshop took place and the national standard, IRAM 17146, was elaborated as well. In this work, the laboratories performance and its temporal evolution is shown from 1997 up to 2002. The suggestions to improve their performance are also included. (author)

  17. Dosimetry methods in boron neutron capture therapy

    Energy Technology Data Exchange (ETDEWEB)

    Gambarini, G.; Artuso, E.; Felisi, M.; Regazzoni, V.; Giove, D. [Universita degli Studi di Milano, Department of Physics, Via Festa del Patrono 7, 20122 Milano (Italy); Agosteo, S.; Barcaglioni, L. [Istituto Nazionale di Fisica Nucleare, Milano (Italy); Campi, F.; Garlati, L. [Politecnico di Milano, Energy Department, Piazza Leonardo Da Vinci 32, 20133 Milano (Italy); De Errico, F. [Universita degli Studi di Pisa, Department of Civil and Industrial Engineering, Lungamo Pacinotti 43, 56126 Pisa (Italy); Borroni, M.; Carrara, M. [Fondazione IRCCS Istituto Nazionale Tumori, Medical Physics Unit, Via Venezian 1, 20133 Milano (Italy); Burian, J.; Klupak, V.; Viererbl, L.; Marek, M. [Research Centre Rez, Department of Neutron Physics, 250-68 Husinec-Rez (Czech Republic)

    2014-08-15

    Dosimetry studies have been carried out at thermal and epithermal columns of Lvr-15 research reactor for investigating the spatial distribution of gamma dose, fast neutron dose and thermal neutron fluence. Two different dosimetry methods, both based on solid state detectors, have been studied and applied and the accuracy and consistency of the results have been inspected. One method is based on Fricke gel dosimeters that are dilute water solutions and have good tissue equivalence for neutrons and also for all the secondary radiations produced by neutron interactions in tissue or water phantoms. Fricke gel dosimeters give the possibility of separating the various dose contributions, i.e. the gamma dose, the fast neutron dose and the dose due to charged particles generated during thermal neutron reactions by isotopes having high cross section, like 10-B. From this last dose, thermal neutron fluence can be obtained by means of the kerma factor. The second method is based on thermoluminescence dosimeters. In particular, the developed method draw advantage from the different heights of the peaks of the glow curve of such phosphors when irradiated with photons or with thermal neutrons. The results show that satisfactory results can be obtained with simple methods, in spite of the complexity of the subject. However, the more suitable dosimeters and principally their utilization and analysis modalities are different for the various neutron beams, mainly depending on the relative intensities of the three components of the neutron field, in particular are different for thermal and epithermal columns. (Author)

  18. Dosimetry methods in boron neutron capture therapy

    International Nuclear Information System (INIS)

    Gambarini, G.; Artuso, E.; Felisi, M.; Regazzoni, V.; Giove, D.; Agosteo, S.; Barcaglioni, L.; Campi, F.; Garlati, L.; De Errico, F.; Borroni, M.; Carrara, M.; Burian, J.; Klupak, V.; Viererbl, L.; Marek, M.

    2014-08-01

    Dosimetry studies have been carried out at thermal and epithermal columns of Lvr-15 research reactor for investigating the spatial distribution of gamma dose, fast neutron dose and thermal neutron fluence. Two different dosimetry methods, both based on solid state detectors, have been studied and applied and the accuracy and consistency of the results have been inspected. One method is based on Fricke gel dosimeters that are dilute water solutions and have good tissue equivalence for neutrons and also for all the secondary radiations produced by neutron interactions in tissue or water phantoms. Fricke gel dosimeters give the possibility of separating the various dose contributions, i.e. the gamma dose, the fast neutron dose and the dose due to charged particles generated during thermal neutron reactions by isotopes having high cross section, like 10-B. From this last dose, thermal neutron fluence can be obtained by means of the kerma factor. The second method is based on thermoluminescence dosimeters. In particular, the developed method draw advantage from the different heights of the peaks of the glow curve of such phosphors when irradiated with photons or with thermal neutrons. The results show that satisfactory results can be obtained with simple methods, in spite of the complexity of the subject. However, the more suitable dosimeters and principally their utilization and analysis modalities are different for the various neutron beams, mainly depending on the relative intensities of the three components of the neutron field, in particular are different for thermal and epithermal columns. (Author)

  19. Acoustic evaluation of polymer gel dosimeters

    International Nuclear Information System (INIS)

    Mather, M.L.; De Deene, Y.; Baldock, C.; Whittaker, A.K.

    2002-01-01

    Advances in radiotherapy treatment techniques such as intensity modulated radiotherapy are placing increasing demands on radiation dosimetry for verification of dose distributions in 3D. In response, polymer gel dosimeters that are capable of recording dose distributions in 3D are currently being developed. Recently, a new technique for evaluation of absorbed dose distributions in these dosimeters using ultrasound was introduced. The current work aims to demonstrate the potential of ultrasound as an evaluation technique for polymer gel dosimeters and to investigate the ultrasound properties of two different dosimeter formulations, PAG and MAGIC gels

  20. Neutron Measurements At Hanford's Plutonium Finishing Plant

    International Nuclear Information System (INIS)

    Conrady, Matthew M.; Berg, Randal K.; Scherpelz, Robert I.; Rathbone, Bruce A.

    2009-01-01

    The Pacific Northwest National Laboratory (PNNL) conducted neutron measurements at Hanford's Plutonium Finishing Plant (PFP). The measurements were performed to evaluate the performance of the Hanford Standard Dosimeter (HSD) and the 8816 TLD component of the Hanford Combination Neutron Dosimeter (HCND) in the neutron fields responsible for worker neutron exposures. For this study, TEPC detectors and multisphere spectrometers were used to measure neutron dose equivalent rate, and multispheres were used to measure average neutron energy. Water-filled phantoms holding Hanford dosimeters were positioned at each measurement location. The phantoms were positioned in the same location where a multisphere measurement was taken and TEPCs were also positioned there. Plant survey meters were also used to measure neutron dose rates at all locations. Three measurement locations were chose near the HC-9B glovebox in room 228A of Building 234-5. The multisphere spectrometers measured average neutron energies in the range of 337 to 555 keV at these locations. Personal dose equivalent, Hp(10)n, as measured by the multisphere and TEPC, ranged from 2.7 to 9.7 mrem/h in the three locations. Effective dose assuming a rotational geometry (EROT) was substantially lower than Hp(10), ranging from 1.3 to 3.6 mrem/h. These values were lower than the reported values from dosimeters exposed on a rotating phantom. Effective dose assuming an AP geometry (EAP) was also substantially lower than Hp(10), ranging from 2.3 to 6.5 mrem/h. These values were lower than the reported values from the dosimeters on slab phantoms. Since the effective dose values were lower than reported values from dosimeters, the dosimeters were shown to be conservative estimates of the protection quantities.

  1. Calculation of neutron fluence-to-dose conversion factors for extremities

    International Nuclear Information System (INIS)

    Stewart, R.D.; Harty, R.; McDonald, J.C.; Tanner, J.E.

    1993-04-01

    The Pacific Northwest Laboratory is developing a standard for the performance testing of personnel extremity dosimeters for the US Department of Energy. Part of this effort requires the calculation of neutron fluence-to-dose conversion factors for finger and wrist extremities. This study focuses on conversion factors for two types of extremity models: namely the polymethyl methacrylate (PMMA) phantom (as specified in the draft standard for performance testing of extremity dosimeters) and more realistic extremity models composed of tissue-and-bone. Calculations for each type of model are based on both bare and D 2 O-moderated 252 Cf sources. The results are then tabulated and compared with whole-body conversion factors. More appropriate energy-averaged quality factors for the extremity models have also been computed from the neutron fluence in 50 equally spaced energy bins with energies from 2.53 x 10 -8 to 15 MeV. Tabulated results show that conversion factors for both types of extremity phantom are 15 to 30% lower than the corresponcung whole-body phantom conversion factors for 252 Cf neutron sources. This difference in extremity and whole-body conversion factors is attributable to the proportionally smaller amount of back-scattering that occurs in the extremity phantoms compared with whole-body phantoms

  2. Development of an alanine dosimeter for gamma dosimetry in mixed environments

    International Nuclear Information System (INIS)

    Vehar, D.W.; Griffin, P.J.

    1992-01-01

    L-αa-Alanine, a nontoxic polycrystalline amino acid, has been investigated for use in high-precision, high-level absorbed-dose measurements in mixed neutron/photon environments such as research and test reactors. The technique is based on the use of electron paramagnetic resonance (EPR) spectroscopy to determine the extent of free radical production in a sample exposed to ionizing radiation, and has been successfully used for photon absorbed-dose measurements at levels exceeding 10 5 Gy with high measurement precision. Application of the technique to mixed environments requires knowledge of the energy-dependent response of the dosimeter for both photons and neutrons. Determination of the dosimeter response to photons is accomplished by irradiations in 60 Co and bremsstrahlung sources and by calculations of energy-dependent photon kerma. Neutron response is determined by calculations in conjunction with CaF 2 :Mn thermoluminescence dosimeters and by calculations of energy-dependent neutron kerma. Several neutron environments are used, including the ACRR and SPR-III reactors

  3. Development of an alanine dosimeter for gamma dosimetry in mixed environments -- Summary of research

    International Nuclear Information System (INIS)

    Vehar, D.W.; Griffin, P.J.

    1994-02-01

    L-α-alanine, a nontoxic polycrystalline amino acid, has been investigated for use in high-precision, high-level absorbed-dose measurements in mixed neutron/photon environments such as research and test reactors. The technique is based on the use of electron paramagnetic resonance spectroscopy to determine the extent of free radical production in a sample exposed to ionizing radiation, and has been successfully used for photon absorbed-dose measurements at levels exceeding 10 5 Gy with high measurement precision. Application of the technique to mixed environments requires knowledge of the energy-dependent response of the dosimeter for both photons and neutrons. Determination of the dosimeter response to photons is accomplished by irradiations in 60 Co and bremsstrahlung sources and by calculations of energy-dependent photon kerma. Neutron response is determined by irradiations in conjunction with CaF 2 :Mn thermoluminescence dosimeters and by calculations of energy-dependent neutron kerma. Several neutron environments are used, including those provided by the Annular Core Research Reactor and Sandia Pulsed Reactor

  4. Personnel Preparation.

    Science.gov (United States)

    Fair, George, Ed.; Stodden, Robert, Ed.

    1981-01-01

    Three articles comprise a section on personnel preparation in vocational education. Articles deal with two inservice programs in career/vocational education for the handicapped and a project to train paraprofessionals to assist special educators in vocational education. (CL)

  5. Development of colored alumilite dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Obara, Kenjiro; Shibanuma, Kiyoshi [Japan Atomic Energy Research Inst., Naka, Ibaraki (Japan). Naka Fusion Research Establishment; Yagi, Toshiaki [Japan Atomic Energy Research Inst., Takasaki, Gunma (Japan). Takasaki Radiation Chemistry Research Establishment; Yokoo, Noriko [Radiation Application Development Association, Tokai, Ibaraki (Japan)

    2003-03-01

    In the ITER (International Thermonuclear Experimental Reactor), in-vessel components such as blanket and divertor, which are installed in the vacuum vessel of the ITER, are maintained by remote handling equipment (RH equipment). The RH equipment for maintenance is operated under sever environmental conditions, such as high temperature (50{approx}100 degC), high gamma-ray radiation ({approx}1 kGy/h) in an atmosphere of inert gas or vacuum; therefore many components of the RH equipment must have a suitable radiation resistance efficiency for long time operation (10{approx}100 MGy). Typical components of the RH equipment have been extensively tested under an intensive gamma-ray radiation. Monitoring of the radiation dose of the components of the RH equipment is essential to control the operation period of the RH equipment considering radiation resistance. However, the maximum measurable radiation dose of the conventional dosimeters, such as ionization chamber, liquid, glass and plastic dosimeters are limited to be approximately 1MGy which is too low to monitor the RH equipment for the ITER. In addition, these conventional dosimeters do not involve sufficient radiation resistance against the high gamma-ray radiation as well as are not easy handling and low cost. Based on the above backgrounds, a new dosimeter with bleaching of an azo group dye to be applied to a radiation monitor has been developed for high gamma-ray radiation use. The colored alumilite dosimeter is composed of the azo group dye (-N=N-) in an anodic oxidation layer of aluminum alloy (Al{sub 2}O{sub 3}). It can monitor the radiation dose by measuring the change of the bleaching of azo dye in the Al{sub 2}O{sub 3} layer due to gamma-ray irradiation. The degree of bleaching is measured as the change of hue (color) and brightness based on the Munsell's colors with a three dimensional universe using spectrophotometer. In the tests, the dependencies such as colors, anodized layer thickness, type of azo

  6. Calibration of personnel monitors by exposure to gamma radiation with energies up to 9 MeV

    International Nuclear Information System (INIS)

    Leao, J.L.B.; Cunha, P.G. da; Diz, R.; Oberhofer, M.

    Occupational exposure with photons of high energy (higher than those from 60 Co) might occur from nuclear reactors and accelerators. Radiation monitors for absorbed dose determination, however, often do not have a wall thickness sufficient to establish electronical equilibrium in that energy range. Using calibration factors determined for lower energies (calibration factors for 60 Co radiation) might cause a significant underestimation of the soft tissue absorbed dose. The calibration factor of the personnel monitors of the Eberline Instrument Co.based on TL dosimeters of LiF is reported. This monitor was calibrated with 9 MeV photons produced by thermal neutrons capture in a Nitarget, at an absorbed dose rate of 125 rads/h at the C.E.N. in France. The results are compared with the corresponding calibration factors for different energies up to 60 Co radiation determined in the IRD (Instituto de Radioprotecao e Dosimetria Rio de Janeiro). (Author) [pt

  7. A PC based thin film dosimeter system

    DEFF Research Database (Denmark)

    Miller, A.; Hargittai, P.; Kovacs, A.

    2000-01-01

    A dosimeter system based on the Riso B3 dosimeter film, an office scanner for use with PC and the associated software is presented. The scanned image is analyzed either with standard software (Paint Shop Pro 5 or Excel) functions or with the computer code "Scanalizer" that allows presentation...

  8. A personal radio-frequency dosimeter with cumulative-dose recording capabilities

    International Nuclear Information System (INIS)

    Rochelle, R.W.; Moore, M.R.; Thomas, R.S.; Ewing, P.D.; Hess, R.A.; Hoffheins, B.S.

    1990-01-01

    The radio-frequency (rf) dosimeter developed by the Oak Ridge National Laboratory is a portable, pocket-sized cumulative-dose recording device designed to detect and record the strengths and durations of electric fields present in the work areas of naval vessels. The device measures an integrated dose and records the electric fields that exceed the permissible levels set by the American National Standards Institute. Features of the rf dosimeter include a frequency range of 30 MHz to 10 GHz and a three-dimensional sensor. Data obtained with the rf dosimeter will be used to determine the ambient field-strength profile for shipboard personnel over an extended time. Readings are acquired and averaged over a 6-min period corresponding to the rise time of the core body temperature. These values are stored for up to 6 months, after which the data are transferred to a computer via the dosimeter's serial port. The rf dosimeter should increase knowledge of the levels of electric fields to which individuals are exposed. 13 refs., 16 figs., 2 tabs

  9. Measurements of SNAC2 area dosimeters placed in different configurations around the PROSPERO reactor and comparison with TRIPOLI-4 calculations

    Energy Technology Data Exchange (ETDEWEB)

    Rousseau, G.; Chambru, L.; Authier, N. [CEA, Centre de Valduc, 21120 Is-sur-Tille, (France)

    2015-07-01

    In the context of criticality accident alarm system tests, several experiments were carried out in 2013 on the PROSPERO reactor to study the response to neutron and gamma of different devices and dosimeters, particularly on the SNAC2 dosimeter. This article presents the results of this criticality dosimeter in different configurations, and compares the experimental measurements with the results of calculation performed with the TRIPOLI-4 Monte-Carlo Neutral Particles transport code. PROSPERO is a metallic critical assembly managed by the Criticality, Neutron Science and Measurement Department located at the French CEA Research Center of Valduc. The core, surrounded by a reflector of depleted uranium, is composed of 2 horizontal cylindrical blocks made of a highly enriched uranium alloy which can be placed in contact, and of 4 depleted uranium control rods which allow the reactor to be driven. This reactor, placed in a cell 10 m x 8 m x 6 m high, with 1.4-meter-thick concrete walls, is used as a fast neutron spectrum source and is operated at stable power level in delayed critical state, which can vary from 3 mW to 3 kW. PROSPERO is extensively used for electronic hardening or to study the effect of the neutrons on various materials. The SNAC2 criticality dosimeter is a zone dosimeter allowing the off line measurement of criticality accident neutron doses. This dosimeter consists of the pile up of seven activation foils embedded into a 23 mm diameter x 21 mm height cadmium container. The activation measurement of each foil, using a gamma spectroscopy technique, gives information about the neutron reaction rates. The SNAC2 software allows the spectrum unfolding from these values, taking into account the hypothesis of a particular spectrum shape, in three components: a Maxwell spectrum component for the thermal range, a 1/E component for the epithermal range, and a Watt spectrum component for the high energy range. Moreover, from the neutron spectrum, the SNAC

  10. Training of reverse propagation neural networks applied to neutron dosimetry

    International Nuclear Information System (INIS)

    Hernandez P, C. F.; Martinez B, M. R.; Leon P, A. A.; Espinoza G, J. G.; Castaneda M, V. H.; Solis S, L. O.; Castaneda M, R.; Ortiz R, M.; Vega C, H. R.; Mendez V, R.; Gallego, E.; De Sousa L, M. A.

    2016-10-01

    Neutron dosimetry is of great importance in radiation protection as aims to provide dosimetric quantities to assess the magnitude of detrimental health effects due to exposure of neutron radiation. To quantify detriment to health is necessary to evaluate the dose received by the occupationally exposed personnel using different detection systems called dosimeters, which have very dependent responses to the energy distribution of neutrons. The neutron detection is a much more complex problem than the detection of charged particles, since it does not carry an electric charge, does not cause direct ionization and has a greater penetration power giving the possibility of interacting with matter in a different way. Because of this, various neutron detection systems have been developed, among which the Bonner spheres spectrometric system stands out due to the advantages that possesses, such as a wide range of energy, high sensitivity and easy operation. However, once obtained the counting rates, the problem lies in the neutron spectrum deconvolution, necessary for the calculation of the doses, using different mathematical methods such as Monte Carlo, maximum entropy, iterative methods among others, which present various difficulties that have motivated the development of new technologies. Nowadays, methods based on artificial intelligence technologies are being used to perform neutron dosimetry, mainly using the theory of artificial neural networks. In these new methods the need for spectrum reconstruction can be eliminated for the calculation of the doses. In this work an artificial neural network or reverse propagation was trained for the calculation of 15 equivalent doses from the counting rates of the Bonner spheres spectrometric system using a set of 7 spheres, one of 2 spheres and two of a single sphere of different sizes, testing different error values until finding the most appropriate. The optimum network topology was obtained through the robust design

  11. Aerated Systems of the Type RH-RCl-Ethanol-Thymolsulphonphthalein Stable Low-Level Chemical Dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Dvornik, I.; Zec, U.; Anic, A.; Ranogajec, F. [Institute Ruder Boskovic, Zagreb, Yugoslavia (Croatia)

    1967-03-15

    The characteristic of dosimeters described in this paper is concerned with the very sensitive colorimetric method of dose evaluation giving a fair sensitivity with low G(HC1). In addition, the systems are thermally stable and simple to manufacture. With photocolorimetric or spectrophotometric evaluation of about 100 rad the dosimetric: error can be as low as 1 rad, or lower. The examined technique of visual colorimetric evaluation at the same dose level gives the combined error of 10-20 rad, and up to {+-} 5 or 10% at 500 rad. Owing to the practically unlimited shelf life of dosimeters and visual colorimeters, and to the very low production costs of both devices, such chemical dosimeters could be of special interest for massive use as personal gamma dosimeters for wide populations, or as dosimeters for gamma and fast neutron dosimetric topography of nuclear accidents. With tetrachloroethylene and iso-octane G(HC1) has been found constant (8.4) for temperatures of between -10 and +35 Degree-Sign C and for dose-rates of between 80 and 80 000 rad/h. The upper dose limit of colorimetric evaluation is about 2000 rad. With other components G(HC1) can be lower and the range extends to higher doses. The colorimetric properties of the systems RH-ethanol-thymolsulphonphthalein, as well as some of the most interesting features of the production procedure, are described. The radiation chemical aspects are discussed briefly. (author)

  12. Personal fast neutrons dosimetry using radiophotoluminescent glass

    International Nuclear Information System (INIS)

    Salem, Y. O.; Nachab, A.; Nourreddine, A.; Roy, C.

    2013-06-01

    In a previous paper we described a new ambient RPL dosimeter that detects fast neutrons in a mixed n-γ field via (n, p) reactions in a polyethylene converter. In the present study, a personal dosimeter is introduced to enable evaluating the individual dose equivalent H p (10) taking into account the albedo. A calibration factor for estimating H p (10) has been determined from the diminishing angular response as the angle of neutron incidence increases to 60 deg from the normal. MCNPX simulations for 241 Am-Be and 252 Cf neutrons, together with a series of monoenergetic neutron beams from 0.144 to 5 MeV, have been used to characterize the dosimeter response, which agrees well with the experimental 241 Am-Be response. (authors)

  13. Development of Thermoluminescence Dosimeter CaSO4:Dy as Personal and Environmental Dosimeters

    International Nuclear Information System (INIS)

    Hasnel Sofyan

    2009-01-01

    Development of personal and environmental dosimeters using material phosphors of CaSO 4 :Dy powder in form capillary glass and disc teflon thermoluminescence (TL) dosimeter have been done. TL dosimeter CaSO 4 :Dy powder used can record dose response less than 0.01 mGy. Fading of TL dosimeter capillary glass after 29 days is 25%. In 1 batch, making of CaSO 4 :Dy powder can obtain 2 groups of dosimeter capillaries with coefficient variance smaller than 10%. This discrepancy caused difference in powder making and reading of the TL dosimeter. TL dosimeter CaSO 4 :Dy teflon disc with dia. 5 mm and 0.8 mm thickness is homogeneous mixture between phosphor powder with dia. 80 to 150 mesh and teflon powder dia. 20 μm. The composition of CaSO 4 :Dy and teflon in TL dosimeter influence sensitivity of the dosimeter. It’ concluded that in order to obtain optimal sensitivity of TL dosimeter, the composition of CaSO 4 :Dy and teflon is 3 and 1 with pressured of disc in 700 MPa. (author).

  14. Scintillation counter based radiation dosimeter

    International Nuclear Information System (INIS)

    Shin, Jeong Hyun

    2009-02-01

    The average human exposure per year is about 240mrem which is come from Radon and human body and terrestrial and cosmic radiation and man-made source. Specially radiation exposure through air from environmental radiation sources is 80mrem/yr(= 0.01mR/hr) which come from Terrestrial and cosmic radiation. Radiation dose is defined as energy deposit/mass. There are two major methods to detect radiation. First method is the energy integration using Air equivalent material like GM counter wall material. Second method is the spectrum to dose conversion method using NaI(Tl), HPGe. These two methods are using generally to detect radiation. But these methods are expensive. So we need new radiation detection method. The research purpose is the development of economical environmental radiation dosimeter. This system consists of Plastic/Inorganic scintillator and Si photo-diode based detector and counting based circuitry. So count rate(cps) can be convert to air exposure rate(R/hr). There are three major advantages in this system. First advantages is no high voltage power supply like GM counter. Second advantage is simple electronics. Simple electronics system can be achieved by Air-equivalent scintillation detector with Al filter for the same detection efficiency vs E curve. From former two advantages, we can know the most important advantages of the this system. Third advantage is economical system. The price of typical GM counter is about $1000. But the price of our system is below $100 because of plastic scintillator and simple electronics. The role of scintillation material is emitting scintillation which is the flash of light produced in certain materials when they absorb ionizing radiation. Plastic scintillator is organic scintillator which is kind of hydrocarbons. The special point are cheap price, large size production(∼ton), moderate light output, fast light emission(ns). And the role of Al filter is equalizing counting efficiency of air and scintillator for

  15. Personnel hazards from medical electron accelerator photoneutrons

    International Nuclear Information System (INIS)

    McCall, R.C.; Jenkins, T.M.; Shore, R.A.; LaRiviere, P.D.

    1979-12-01

    For medical accelerators, neutron penetration through the room entry door is the major personnel hazard. Most therapy accelerator rooms are designed with at least a rudimentary maze to avoid the use of massive doors. Often, however, the maze may be similar to those shown in scale outline drawings of some medical electron accelerator rooms where the authors have made neutron measurements outside the doors which were of different thicknesses and compositions. The results are tabulated. It should be noted that there can be significant dose equivalents (H) at the door when a maze is inadequate, and that all three components - fast neutron, thermal neutron, and neutron capture γ rays - can be equally important

  16. Survey of neutrons inside the containment of a pressurized water reactor

    International Nuclear Information System (INIS)

    Hankins, D.E; Griffith, R.V.

    1978-01-01

    A neutron survey was made inside the containment of the Farley Nuclear Plant, Alabama Power and Light Company, Dothan, Alabama, in November 1977. The survey was made to determine the spectra of leakage neutrons and to evaluate the accuracy of albedo neutron dosimeters and a 9-in.-diameter sphere rem meter. The survey also covered variations in the neutron spectra, the ratio of gamma-to-neutron dose rates, and the thermal neutron component of the neutron dose

  17. Dosimetric properties of the pocket alarm dosimeter type Graetz TDW 10/20

    International Nuclear Information System (INIS)

    Hauser, M.; Burgkhardt, B.; Piesch, E.

    1980-08-01

    In personnel monitoring pocket dosimeters with build-in alarm devices are increasingly in use. The report presents results of a test performed at Karlsruhe. The properties investigated are above all linearity and reproducibility of the dose reading as well as of the acustic alarm indication, dependence on the dose reading, the photon energy, the direction of the radiation incidence, the dose rate, the temperature and operational characteristic of the batteries. (orig./HP) [de

  18. Evaluating secondary neutron doses of a refined shielded design for a medical cyclotron using the TLD approach

    International Nuclear Information System (INIS)

    Lin, Jye-Bin; Tseng, Hsien-Chun; Liu, Wen-Shan; Lin, Ding-Bang; Hsieh, Teng-San; Chen, Chien-Yi

    2013-01-01

    An increasing number of cyclotrons at medical centers in Taiwan have been installed to generate radiopharmaceutical products. An operating cyclotron generates immense amounts of secondary neutrons from reactions such the 18 O(p, n) 18 F, used in the production of FDG. This intense radiation can be hazardous to public health, particularly to medical personnel. To increase the yield of 18 F-FDG from 4200 GBq in 2005 to 48,600 GBq in 2011, Chung Shan Medical University Hospital (CSMUH) has prolonged irradiation time without changing the target or target current to meet requirements regarding the production 18 F. The CSMUH has redesigned the CTI Radioisotope Delivery System shield. The lack of data for a possible secondary neutron doses has increased due to newly designed cyclotron rooms. This work aims to evaluate secondary neutron doses at a CTI cyclotron center using a thermoluminescent dosimeter (TLD-600). Two-dimensional neutron doses were mapped and indicated that neutron doses were high as neutrons leaked through self-shielded blocks and through the L-shaped concrete shield in vault rooms. These neutron doses varied markedly among locations close to the H 2 18 O target. The Monte Carlo simulation and minimum detectable dose are also discussed and demonstrated the reliability of using the TLD-600 approach. Findings can be adopted by medical centers to identify radioactive hot spots and develop radiation protection. - Highlights: • Neutron doses were verified using TLD approach. • Neutron doses were increased at cyclotron centers. • Revised L-shaped shield suppresses effectively the neutrons. • Neutron dose can be attenuated to 1.13×10 6 %

  19. Performing personnel dosimetry investigations and records quality assurance

    International Nuclear Information System (INIS)

    Perle, S.C.

    2002-01-01

    Radiation Safety Officers (RSOs) sometimes face situations in which personnel dosimetry estimates are required after dosimeters issued to radiation workers (film or TLD badges, extremity dosimeters, etc.) are lost or damaged before processing. This article was prepared to help those involved with personnel dosimetry investigations became aquatinted with this process. A factor that contributes to the anxiety of those unfamiliar with dosimetry investigations is the lack of published guidance available in this subject. More printed resources are needed to help radiation safety professionals familiarize themselves and understand personnel dosimetry investigations. Topics discussed in this presentation include the justification of performing dosimetry investigations, recommendations on how to perform them and the advantages of performing such investigations

  20. Epoxy encapsulant as serendipitous dosimeters during radiological/nuclear events

    Energy Technology Data Exchange (ETDEWEB)

    Barkyoumb, J.H. [Carderock Division, Naval Surface Warfare Center, 9500 MacArthur Blvd., West Bethesda, MD 20817-5700 (United States)], E-mail: jhbarky@earthlink.net; Mathur, V.K. [Carderock Division, Naval Surface Warfare Center, 9500 MacArthur Blvd., West Bethesda, MD 20817-5700 (United States)

    2008-02-15

    The radiation response of a smart chip (embedded integrated circuit) module has been reported earlier using the technique of optically stimulated luminescence (OSL). It was found that a smart chip module could be used to evaluate the personnel exposure in the accident dosimetry range. Through subsequent experiments, the radiation sensitivity of the chip module was traced to the epoxy encapsulant provided to protect the chip from the environment and physical damage and that the radiation sensitivity of the epoxy is due to the silica used as the 'filler' for controlling the thixotropic properties of the epoxy used for 'glob top' or 'dam-and-fill' encapsulation. It is desirable to retain the ability to use the smart chip as an accident dosimeter without requiring a modification of standard manufacturing process for which an infrastructure already exists to avoid additional costs. For this reason, we have investigated commercially available filled and unfilled epoxies both as received from the manufacturer and compared their response with epoxies to which commercial fillers are added. In this work we investigate the OSL response of various epoxies commonly used for potting of electronic circuits with and without various filler materials for their potential to be used as a casualty dosimeter in the exposure range of 0.5-10 Gy.

  1. Analysis of personnel monitoring control card data

    International Nuclear Information System (INIS)

    Ande, C.D.; Sneha, C.; Madhumita, B.; Bakshi, A.K.; Datta, D.

    2018-01-01

    In India, personnel monitoring of radiation workers for X-, beta- and gamma- radiation is carried out using a thermoluminescence dosemeter (TLD) system based on CaSO 4 :Dy Teflon TLD disc. A large number of radiation workers get very low occupational doses and their doses are reported as zero since it is not above detectable limits. Therefore, the detection of low levels of occupational dose over and above the natural background assumes great importance. In the present system, the estimation of the background dose is achieved by use of control dosemeters. An analysis of the readings of the control dosimeters sent to various institutions was carried out to arrive at conclusions regarding the validity of the use of control dosimeter

  2. Development of a TL personal dosimeter identifiable PA exposure, and comparison with commercial TL dosimeters

    International Nuclear Information System (INIS)

    Kwon, J.W.; Kim, H.K.; Lee, J.K.; Kim, J.L.

    2004-01-01

    A single-dosimeter worn on the anterior surface of the body of a worker was found to significantly underestimate the effective dose to the worker when the radiation comes from the back. Several researchers suggested that this sort of underestimation can be corrected to a certain extent by using an extra dosimeter on the back. However, use of multiple dosimeters also has disadvantages such as complication in control or incurrence of extra cost. Instead of the common multi-dosimeter approach, in this study, a single dosimeter introducing asymmetric filters which enabled to identify PA exposure was designed, and its dose evaluation algorithm for AP-PA mixed radiation fields was established. A prototype TL personal dosimeter was designed and constructed. The Monte Carlo simulations were utilized in the design process and verified by experiments. The dosimeter and algorithm were applicable to photon radiation having an effective energy beyond 100 keV in AP-PA mixed radiation fields. A simplified performance test based on ANSI N13.11 showed satisfactory results. Considering that the requirements of the International Electrotechnical Commission (IEC) and the American National Standards Institute (ANSI) with regard to the dosimeter on angular dependency is reinforced, the dosimeter and the dose evaluation algorithm developed in this study provides a useful approach in practical personal dosimetry against inhomogeneous high energy radiation fields. (author)

  3. Personnel radiation monitoring by thermoluminescence dosimetry (1995-96)

    International Nuclear Information System (INIS)

    Daw Mi Cho Cho; Daw Yi Yi Khin; Daw San San; U Maung Maung Tin; Daw Hla Hla Win

    2001-01-01

    Personnel radiation monitoring which is the dose assessment of individual doses from external radiation received by radiation workers has been carried out by Thermoluminescence Dosimetry system consisting of a Vinten Toledo TLD reader, LiF dosimeters and associated equipment. The exposed TLD dosimeters were measured by TLD reader and the dose evaluation and dose registration were done on personal computer. Due to the records of 1995-96, most of the radiation workers complied with the permissible dose recommended by IAEA and ICRP 60. (author)

  4. Miniature Active Space Radiation Dosimeter, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Space Micro will extend our Phase I R&D to develop a family of miniature, active space radiation dosimeters/particle counters, with a focus on biological/manned...

  5. The shelf life of dyed polymethylmethacrylate dosimeters

    International Nuclear Information System (INIS)

    Bett, R.; Watts, M.F.; Plested, M.E.

    2002-01-01

    The long-term stability of the radiation response of Harwell Red 4034 and Amber 3042 Perspex Dosimeters has been monitored for more than 15 years, and the resulting data used in the justification of their shelf-life specifications

  6. An improved dosimeter having constant flow pump

    International Nuclear Information System (INIS)

    Baker, W.B.

    1980-01-01

    A dosemeter designed for individual use which can be used to monitor toxic radon gas and toxic related products of radon gas in mines and which incorporates a constant air stream flowing through the dosimeter is described. (U.K.)

  7. Floating Gate CMOS Dosimeter With Frequency Output

    Science.gov (United States)

    Garcia-Moreno, E.; Isern, E.; Roca, M.; Picos, R.; Font, J.; Cesari, J.; Pineda, A.

    2012-04-01

    This paper presents a gamma radiation dosimeter based on a floating gate sensor. The sensor is coupled with a signal processing circuitry, which furnishes a square wave output signal, the frequency of which depends on the total dose. Like any other floating gate dosimeter, it exhibits zero bias operation and reprogramming capabilities. The dosimeter has been designed in a standard 0.6 m CMOS technology. The whole dosimeter occupies a silicon area of 450 m250 m. The initial sensitivity to a radiation dose is Hz/rad, and to temperature and supply voltage is kHz/°C and 0.067 kHz/mV, respectively. The lowest detectable dose is less than 1 rad.

  8. Perfection of the individual photographic emulsion dosimeter

    International Nuclear Information System (INIS)

    Soudain, G.

    1960-01-01

    A photographic dosimeter making possible the measurement of γ radiation doses of from 10 mr up to 800 r by means of 3 emulsion bands of varying sensitivity stuck to the same support is described. The dosimeter has also a zone for marking and a test film insensitive to radiation. This requires a photometric measurement by diffuse reflection an d makes it possible to measure doses with an accuracy of 20 per cent. (author) [fr

  9. EVIDOS: Individual dosimetry in mixed neutron and photon radiation fields

    International Nuclear Information System (INIS)

    Vanhavere, F.

    2006-01-01

    The EVIDOS project (partly funded by the European Commission RTD Programme: Nuclear Energy, Euratom Framework Programme V, 1998-2002, Contract No FIKR-CT-2001-00175) aimed at improving individual monitoring in mixed neutron-photon radiation fields by evaluating the performance of routine and novel personal dosimeters for mixed radiation, and by giving guidelines for deriving sufficiently accurate values of personal dose equivalent from the readings of area survey instruments and dosimeters. The main objective of EVIDOS was to evaluate different methods for individual dosimetry in mixed neutron-photon work-places in nuclear industry. This implied a determination of the capabilities and limitations of personal dosimeters and the establishment of methods to enable sufficiently accurate values of personal dose equivalent from spectrometers, area survey instruments and routine personal dosimeters. Also novel electronic personal dosimeters were investigated. To this end spectrometric and dosimetric investigations in selected representative workplaces in nuclear industry where workers can receive significant neutron doses were performed. As part of this project, a number of tasks were executed, in particular: (1) the determination of the energy and direction distribution of the neutron fluence; (2) the derivation of the (conventionally true) values of radiation protection quantities; (3) the determination of the readings of routine and innovative personal dosimeters and of area monitors; and (4) the comparison between dosimeter readings and values of the radiation protection quantities

  10. A fibre optic dosimeter customised for brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Suchowerska, N. [Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, NSW 2050 (Australia); School of Physics, University of Sydney, NSW 2006 (Australia)], E-mail: Natalka@email.cs.nsw.gov.au; Lambert, J.; Nakano, T. [Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, NSW 2050 (Australia); School of Physics, University of Sydney, NSW 2006 (Australia); Law, S. [School of Physics, University of Sydney, NSW 2006 (Australia); Optical Fibre Technology Centre, University of Sydney, 206 National Innovation Centre, Australian Technology Park, Eveleigh, NSW 1430 (Australia); Elsey, J. [Bandwidth Foundry Pty Ltd, Australian Technology Park, NSW, 1430 (Australia); McKenzie, D.R. [School of Physics, University of Sydney, NSW 2006 (Australia)

    2007-04-15

    In-vivo dosimetry for brachytherapy cancer treatment requires a small dosimeter with a real time readout capability that can be inserted into the patient to determine the dose to critical organs. Fibre optic scintillation dosimeters, consisting of a plastic scintillator coupled to an optical fibre, are a promising dosimeter for this application. We have implemented specific design features to optimise the performance of the dosimeter for specific in-vivo dosimetry during brachytherapy. Two sizes of the BrachyFOD{sup TM} scintillation dosimeter have been developed, with external diameters of approximately 2 and 1 mm. We have determined their important dosimetric characteristics (depth dose relation, angular dependence, energy dependence). We have shown that the background signal created by Cerenkov and fibre fluorescence does not significantly affect the performance in most clinical geometries. The dosimeter design enables readout at less than 0.5 s intervals. The clinical demands of real time in-vivo brachytherapy dosimetry can uniquely be satisfied by the BrachyFOD{sup TM}.

  11. Polymer gel dosimeter based on itaconic acid

    International Nuclear Information System (INIS)

    Mattea, Facundo; Chacón, David

    2015-01-01

    A new polymeric dosimeter based on itaconic acid and N, N’-methylenebisacrylamide was studied. The preparation method, compositions of monomer and crosslinking agent and the presence of oxygen in the dosimetric system were analyzed. The resulting materials were irradiated with an X-ray tube at 158 cGy/min, 226 cGy min and 298 cGy/min with doses up to 1000 Gy. The dosimeters presented a linear response in the dose range 75–1000 Gy, sensitivities of 0.037 1/Gy at 298 cGy/min and an increase in the sensitivity with lower dose rates. One of the most relevant outcomes in this study was obtaining different monomer to crosslinker inclusion in the formed gel for the dosimeters where oxygen was purged during the preparation method. This effect has not been reported in other typical dosimeters and could be attributed to the large differences in the reactivity among these species. - Highlights: • A novel polymer gel dosimeters based on itaconic acid is presented and characterized. • The typical linear trend of the dose behavior in a specific dose range was found. • Different gel structures were formed when oxygen and an antioxidant were present. • Absorbed dose is univocally correlated with optic absorbance and Raman spectroscopy. • Itaconic acid appears as a reliable radiation dosimeter that may be further improved.

  12. Comparative study of some new EPR dosimeters

    International Nuclear Information System (INIS)

    Alzimami, K.S.; Maghraby, Ahmed M.; Bradley, D.A.

    2014-01-01

    Investigations have been made of four new radiation dosimetry EPR candidates from the same family of materials: sulfamic acid, sulfanillic acid, homotaurine, and taurine. Mass energy attenuation coefficients, mass stopping power values and the time dependence of the radiation induced radicals are compared. Also investigated are the microwave saturation behavior and the effect of applied modulation amplitude on both peak-to-peak line width (W PP ) and peak-to-peak signal height (H PP ). The dosimeters are characterized by simple spectra and stable radiation-induced radicals over reasonable durations, especially in taurine dosimeters. Sulfamic acid dosimeters possessed the highest sensitivity followed by taurine and homotaurine and sulfanillic. - Highlights: ► Several EPR dosimeters were suggested based on SO 3 − radical. ► Taurine, homotaurine, sulfanilic, and sulfamic acid all possess simple EPR spectra. ► Dosimeters were compared to each other in terms of the dosimetric point of view. ► Energy dependence curves of the selected dosimeters were compared to eachother

  13. DNA adducts as molecular dosimeters

    International Nuclear Information System (INIS)

    Lucier, G.W.

    1990-01-01

    There is compelling evidence that DNA adducts play an important role in the actions of many pulmonary carcinogens. During the last ten years sensitive methods (antibodies and 32 P-postlabeling) have been developed that permit detection of DNA adducts in tissues of animals or humans exposed to low levels of some genotoxic carcinogens. This capability has led to approaches designed to more reliably estimate the shape of the dose-response curve in the low dose region for a few carcinogens. Moreover, dosimetry comparisions can, in some cases, be made between animals and humans which help in judging the adequacy of animal models for human risk assessments. There are several points that need to be considered in the evaluation of DNA adducts as a molecular dosimeter. For example, DNA adduct formation is only one of many events that are needed for tumor development and some potent carcinogens do not form DNA adducts; i.e., TCDD. Other issues that need to be considered are DNA adduct heterogeneity, DNA repair, relationship of DNA adducts to somatic mutation and cell specificity in DNA adduct formation and persistence. Molecular epidemiology studies often require quantitation of adducts in cells such as lymphocytes which may or may not be reliable surrogates for adduct concentrations in target issues. In summary, accurate quantitation of low levels of DNA adducts may provide data useful in species to species extrapolation of risk including the development of more meaningful human monitoring programs

  14. Superheated drop, open-quotes Bubbleclose quotes, dosimeters

    International Nuclear Information System (INIS)

    Harper, M.J.; Lindler, K.W.; Nelson, M.E.; Johnson, T.L.; Jones, C.R.; Rabovsky, J.L.; Rao, N.; Kerschner, H.F.; Reil, G.K.; Schwartz, R.B.

    1991-01-01

    Superheated Drop Dosimeters (SDD) offer a sensitive, immediate measure of the neutron dose equivalent, but their dynamic range is limited and their response varies with temperature, pressure, and vibration. They contain thousands of superheated liquid drops in a stabilizing matrix. High linear energy transfer (LET) radiation triggers vaporization of the drops into visible bubbles. If the matrix is a liquid, the bubbles slowly rise, and the number present indicates the dose rate. Dose may be measured by displacement of the matrix, or by counting the sounds of vaporization. If the matrix is a gel, the bubbles are fixed, and their number is proportional to the dose equivalent. Our research has focused on modeling and elimination of the environmental response, extension of the dynamic range, and tests and evaluations of prototype devices

  15. Composite Resin Dosimeters: A New Concept and Design for a Fibrous Color Dosimeter.

    Science.gov (United States)

    Kinashi, Kenji; Iwata, Takato; Tsuchida, Hayato; Sakai, Wataru; Tsutsumi, Naoto

    2018-04-11

    Polystyrene (PS)-based composite microfibers combined with a photochromic spiropyran dye, 1,3,3-trimethylindolino-6'-nitrobenzopyrylospiran (6-nitro BIPS), and a photostimulable phosphor, europium-doped barium fluorochloride (BaFCl:Eu 2+ ), were developed for the detection of X-ray exposure doses on the order of approximately 1 Gy. To produce the PS-based composite microfibers, we employed a forcespinning method that embeds a high concentration of phosphor in PS in a safe, inexpensive, and simple procedure. On the basis of the optimization of the forcespinning process, fibrous color dosimeters with a high radiation dose sensitivity of 1.2-4.4 Gy were fabricated. The color of the dosimeters was found to transition from white to blue in response to X-ray exposure. The optimized fibrous color dosimeter, made from a solution having a PS/6-nitro BIPS/BaFCl:Eu 2+ /C 2 Cl 4 ratio of 7.0/0.21/28.0/28.0 (wt %) and produced with a 290 mm distance between the needle and collectors, a 0.34 mm 23 G needle nozzle, and a spinneret rotational rate of 3000 rpm, exhibited sensitivity to a dose as low as 1.2 Gy. To realize practical applications, we manufactured the optimized fibrous color dosimeter into a clothlike color dosimeter. The clothlike color dosimeter was mounted on a stuffed bear, and its coloring behavior was demonstrated upon X-ray exposure. After exposure with X-ray, a blue colored and shaped in the form of the letter "[Formula: see text]" clearly appeared on the surface of the clothlike color dosimeter. The proposed fibrous color dosimeters having excellent workability will be an unprecedented dosimetry and contributed to all industries utilizing radiation dosimeters. This new fibrous "composite resin dosimeter" should be able to replace traditional, wearable, and individual radiation dose monitoring devices, such as film badges.

  16. Solid thermoluminescent dosemeter of sodium tetraborate and brazilian fluoride sensitive to thermal neutrons

    International Nuclear Information System (INIS)

    Fratin, L.

    1988-01-01

    The techniques of compacting sodium tetraborate and natural fluoride mixtures were studied in this work, with the aim of producing a solid dosimeter sensitive to thermal neutrons. The production procedure involves the vitrification of the sodium tetraborate, the grinding, mixture, cold pressing and the sinterization of the pellets. A special arrangement was built for irradiation where paraffin was used as moderator for neutrons from a 241 Am-Be source. Two different mass ratios of sodium tetraborate and flourite showed a linear thermoluminescent response to the neutron fluence in the range of 1.0 to 7.0 x 10 8 n (sub)tcm -2 . Solid dosimeters, manufactured from natural fluorite and sodium chloride, showed a response to gamma radiation similar to the response of the dosimeters sensitive to neutrons. These dosimeters are need to identify the proportion of thermoluminescent response due to gamma radiation present in a neutron field. (author) [pt

  17. Characterization of TL dosimeters for determination of the gamma component in a mixed n+γ radiation field

    International Nuclear Information System (INIS)

    Ranogajec-Komor, M.; Miljanic, S.; Ferek, S.; Dvornik, I.; Osvay, M.

    1996-01-01

    In the International Intercomparison of the Criticality Accident Dosimetry Systems organized by the Commission of European Communities at SILENE Reactor in Valduc, France, 1993, the Ruder Boskovic Institute (RBI) measured the total neutron and gamma tissue absorbed dose (D n+γ ) at the body surface irrespective of neutron and gamma energy spectra variations using the chemical dosimeters DL-M4. For deriving the neutron dose i.e recoil dose, D n , from the differences D n = D n+γ - D t , the total gamma dose (D tγ ) has to be measured with highest accuracy. The determination of the gamma dose in a mixed field is complicated because TL dosimeters are sensitive both to neutrons and gammas. Besides, the radiation doses and energy spectra vary because of scattering and absorption in the body or phantom. Therefore dosimeters with different sensitivities, energy dependences and encapsulations have to be used. In this paper only the study of some characteristics of various TL detectors, such as sensitivity, linearity, supralinearity and fading, for measurement of the gamma component are described. These investigations were carried out in RBI before and after the Valduc intercomparison experiments. The encapsulations, TL response corrections for thermal and fast neutron effects as well as the discussion of Valduc results will be published later

  18. Blood proteins as carcinogen dosimeters

    International Nuclear Information System (INIS)

    Tannenbaum, S.R.; Skipper, P.L.

    1986-01-01

    The problem of quantifying exposure to genotoxins in a given individual represents a formidable challenge. In this paper methods which rely on the covalent binding of carcinogens and their metabolites to blood proteins are described. That carcinogens interact with proteins as well as with DNA has been established, although whether protein-carcinogen adducts can result in genetic damage has not been established. It has been shown, however, that the amount of a protein carcinogen adduct formed may be used as a quantitative measure of exposure to a carcinogen. Such a measure presumably is reflective of the absorption, metabolism, and excretion of the compound in an exposed individual. Protein adduction may reflect exposure in a time-frame of weeks to months. Thus, protein adduct measurement is a form of human chemical dosimetry. Hemoglobin and albumin are promising candidates for such dosimeters. Hemoglobin has a lifetime of about 120 days in humans; thus, circulating levels of carcinogen-modified hemoglobin will reflect the level of carcinogen exposure during a period of nearly four months. It also possesses some metabolic competence, particularly, the ability to oxidize aromatic hydroxylamines to nitroso compounds which react quite efficiently with sulfhydryl groups. Albumin has a half-life of 20 to 25 days in man. This protein does not possess metabolic capacity other than, perhaps, some esterase activity. In contrast to hemoglobin, though, it is not protected by the erythrocyte membrane and might be the target for a greater number of carcinogens. It is present and is synthesized in the same cells in which the reactive metabolic intermediates of carcinogens are mostly formed - the hepatocytes. Also, albumin has a number of high-affinity binding sites for a broad spectrum of xenobiotics and endobiotics. 25 refs., 1 tab

  19. Quality assurance of the manual OSLD Badge based personnel monitoring system

    International Nuclear Information System (INIS)

    Kumar, Munish; Gaonkar, U.P.; Koul, D.K.; Datta, D.; Singh, S.K.; Rakesh, R.B.; Kulkarni, M.S.; Ratna, P.

    2018-01-01

    Optically stimulated luminescence (OSL) based technique is widely used for monitoring of radiation workers exposed to x, gamma and beta radiations. Features such as multiple readout, short processing time, better beta response, absence of infra red signal (no heating), flexibility in the dosimeter design using binders or substrates having low melting point are associated advantages. Presently α-Al 2 O 3 :C and BeO based dosimeters are popularly used for personnel monitoring applications worldwide. The present paper describe results of the external quality control check (QAC) tests performed for the present OSLD badge based personnel monitoring system. The results were analyzed as per ANSI criteria and Trumpet curve methods

  20. Training of reverse propagation neural networks applied to neutron dosimetry; Entrenamiento de redes neuronales de propagacion inversa aplicadas a la dosimetria de neutrones

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez P, C. F.; Martinez B, M. R.; Leon P, A. A.; Espinoza G, J. G.; Castaneda M, V. H.; Solis S, L. O.; Castaneda M, R.; Ortiz R, M.; Vega C, H. R. [Universidad Autonoma de Zacatecas, Av. Ramon Lopez Velarde 801, Col. Centro, 98000 Zacatecas, Zac. (Mexico); Mendez V, R. [Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas, Laboratorio de Patrones Neutronicos, Av. Complutense 22, 28040 Madrid (Spain); Gallego, E. [Universidad Politecnica de Madrid, Departamento de Ingenieria Nuclear, ETSI Industriales, Jose Gutierrez Abascal 2, 28006 Madrid (Spain); De Sousa L, M. A. [Centro de Desenvolvimento da Tecnologia Nuclear / CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Pampulha, Belo Horizonte, Minas Gerais (Brazil)

    2016-10-15

    Neutron dosimetry is of great importance in radiation protection as aims to provide dosimetric quantities to assess the magnitude of detrimental health effects due to exposure of neutron radiation. To quantify detriment to health is necessary to evaluate the dose received by the occupationally exposed personnel using different detection systems called dosimeters, which have very dependent responses to the energy distribution of neutrons. The neutron detection is a much more complex problem than the detection of charged particles, since it does not carry an electric charge, does not cause direct ionization and has a greater penetration power giving the possibility of interacting with matter in a different way. Because of this, various neutron detection systems have been developed, among which the Bonner spheres spectrometric system stands out due to the advantages that possesses, such as a wide range of energy, high sensitivity and easy operation. However, once obtained the counting rates, the problem lies in the neutron spectrum deconvolution, necessary for the calculation of the doses, using different mathematical methods such as Monte Carlo, maximum entropy, iterative methods among others, which present various difficulties that have motivated the development of new technologies. Nowadays, methods based on artificial intelligence technologies are being used to perform neutron dosimetry, mainly using the theory of artificial neural networks. In these new methods the need for spectrum reconstruction can be eliminated for the calculation of the doses. In this work an artificial neural network or reverse propagation was trained for the calculation of 15 equivalent doses from the counting rates of the Bonner spheres spectrometric system using a set of 7 spheres, one of 2 spheres and two of a single sphere of different sizes, testing different error values until finding the most appropriate. The optimum network topology was obtained through the robust design

  1. Time-dependent, low-dose reporting limit for dosimeters that are taken home at the end of the workday

    International Nuclear Information System (INIS)

    Sonder, E.

    1994-01-01

    In routine personnel dosimetry, it is usual to report doses only where th occupational dose (measured dose with background subtracted) is greater than previously determined reporting limit. The reporting limit, although se administratively, should be justified by an assessment of the errors inherent in th personnel and background dose measurements, and estimates of the probability that a zero exposure will yield a dosimeter response equal to the reporting limit. For background subtraction and reporting limits, it was realized that the source of low-dose uncertainty is very much dependent on whether dosimeters are held in racks at the work site or remain with the employee. The External dosimetry Program for the DOE facilities at Oak Ridge, Tennessee, is operated such that dosimeters are taken home by the employee at the end of each workday. This report is a summary measurements of background radiation in a variety of home locations, and calculations of the effect of the variation of this background on the uncertainty low-level occupational dose and on the reporting level. When dosimeters are stored at a given location (e.g., the facility gate), it theoretically possible to determine the background there to any desired accuracy; the errors in occupational dose are then those due to measurement noise and dosimeter calibration. However, when different dosimeters are stored in different homes, th difference in background between a particular location and the average for a locations appears as an added uncertainty in the occupational dose. Since this difference is not random, but fixed for a given location, the error due to this difference increases linearly with assignment time. From the background measurements, time-dependent values of the standard deviation of occupational do and critical levels are derived and used to define an expression for the reporting limit

  2. Automated Calibration of Dosimeters for Diagnostic Radiology

    International Nuclear Information System (INIS)

    Romero Acosta, A.; Gutierrez Lores, S.

    2015-01-01

    Calibration of dosimeters for diagnostic radiology includes current and charge measurements, which are often repetitive. However, these measurements are usually done using modern electrometers, which are equipped with an RS-232 interface that enables instrument control from a computer. This paper presents an automated system aimed to the measurements for the calibration of dosimeters used in diagnostic radiology. A software application was developed, in order to achieve the acquisition of the electric charge readings, measured values of the monitor chamber, calculation of the calibration coefficient and issue of a calibration certificate. A primary data record file is filled and stored in the computer hard disk. The calibration method used was calibration by substitution. With this system, a better control over the calibration process is achieved and the need for human intervention is reduced. the automated system will be used in the calibration of dosimeters for diagnostic radiology at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene. (Author)

  3. Water equivalence of polymer gel dosimeters

    International Nuclear Information System (INIS)

    Sellakumar, P.; James Jebaseelan Samuel, E.; Supe, Sanjay S.

    2007-01-01

    To evaluate the water equivalence and radiation transport properties of polymer gel dosimeters over the wide range of photon and electron energies 14 different types of polymer gels were considered. Their water equivalence was evaluated in terms of effective atomic number (Z eff ), electron density (ρ e ), photon mass attenuation coefficient (μ/ρ), photon mass energy absorption coefficient (μ en /ρ) and total stopping power (S/ρ) tot of electrons using the XCOM and the ESTAR database. The study showed that the effective atomic number of polymer gels were very close ( en /ρ for all polymer gels were in close agreement ( tot of electrons in polymer gel dosimeters were within 1% agreement with that of water. From the study we conclude that at lower energy (<80keV) the polymer gel dosimeters cannot be considered water equivalent and study has to be carried out before using the polymer gel for clinical application

  4. Fiber-optic dosimeters for radiation therapy

    Science.gov (United States)

    Li, Enbang; Archer, James

    2017-10-01

    According to the figures provided by the World Health Organization, cancer is a leading cause of death worldwide, accounting for 8.8 million deaths in 2015. Radiation therapy, which uses x-rays to destroy or injure cancer cells, has become one of the most important modalities to treat the primary cancer or advanced cancer. The newly developed microbeam radiation therapy (MRT), which uses highly collimated, quasi-parallel arrays of x-ray microbeams (typically 50 μm wide and separated by 400 μm) produced by synchrotron sources, represents a new paradigm in radiotherapy and has shown great promise in pre-clinical studies on different animal models. Measurements of the absorbed dose distribution of microbeams are vitally important for clinical acceptance of MRT and for developing quality assurance systems for MRT, hence are a challenging and important task for radiation dosimetry. On the other hand, during the traditional LINAC based radiotherapy and breast cancer brachytherapy, skin dose measurements and treatment planning also require a high spatial resolution, tissue equivalent, on-line dosimeter that is both economical and highly reliable. Such a dosimeter currently does not exist and remains a challenge in the development of radiation dosimetry. High resolution, water equivalent, optical and passive x-ray dosimeters have been developed and constructed by using plastic scintillators and optical fibers. The dosimeters have peak edge-on spatial resolutions ranging from 50 to 500 microns in one dimension, with a 10 micron resolution dosimeter under development. The developed fiber-optic dosimeters have been test with both LINAC and synchrotron x-ray beams. This work demonstrates that water-equivalent and high spatial resolution radiation detection can be achieved with scintillators and optical fiber systems. Among other advantages, the developed fiber-optic probes are also passive, energy independent, and radiation hard.

  5. Diffusion measurement in ferrous infused gel dosimeters

    International Nuclear Information System (INIS)

    Zahmatkesh, M. H.; Healy, B. J.

    2003-01-01

    Background: The compositions of Ferrous sulphate, Agarose and Xylenol orange dye and Ferrous sulphate, Gelatin and Xylenol orange dye in solution of distilled water and sulphuric acid are two tissue-equivalent gel dosimeters. Ionizing radiation causes oxidation of Fe 2+ ion to Fe 3+ ions which diffuse through the gel matrix and blur the image of absorbed dose over a period of hours after irradiation. Materials and methods: 25 m M sulphuric acid, 0.4 mm ferrous ammonium sulphate, 0.2 mm xylenol orange dye and 1% by weight agarose in distilled water named Agarose and Xylenol orange dye and 0.1 mm ferrous ammonium sulphate, 0.1 mm xylenol orange dye, 50 mm sulphuric acid and 5% by weight gelatin in distilled water named Gelatin and Xylenol orange dye are used as two gel dosimeters. All chemicals were supplied by Sigma Ald ridge Company, Germany. The gels were poured in Perspex casts and were irradiated to a beam of X ray from linear accelerators or X ray machine. Results: In this study diffusion coefficients of Agarose and Xylenol orange dye and Gelatin and Xylenol orange dye dosimeters have been measured through a computer program for different temperature. The ferric ion diffusion coefficient (D) for the Agarose and Xylenol orange dye and Gelatin and Xylenol orange dye dosimeters were measured as (1.19.±0.03) x 10 -2 cm 2 .hr -1 and (0.83±0.03) x 10 -2 cm 2 .hr -1 respectively at room temperature. Conclusion: For both dosimeters the diffusion coefficients decreased with gel storage temperatures down to 6 d ig C . Gelatin and Xylenol orange dye dosimeters have advantage of lower diffusion coefficient for a specified temperature

  6. Development and physical characteristics of a novel compound radiophotoluminescent glass dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Hsu, S.-M. [Department of Biomedical Imaging and Radiological Sciences, National Yang-Ming University, 155, Li-Nong Street Sec. 2, Pei-tou, Taipei 112, Taiwan (China)], E-mail: d49220003@ym.edu.tw; Yang, H.-W. [Department of Material Science and Engineering, National United University, No. 1, Lien Da, Miao-Li 360, Taiwan (China); Huang, David Y.C. [Facutly of Memorial Sloan-Kettering Cancer Center at Mercy Medical Center, 1000 N, Village Avenue, Rockville Centre, NY (United States); Hsu, W.-L.; Lu, C.-C. [Department of Material Science and Engineering, National United University, No. 1, Lien Da, Miao-Li 360, Taiwan (China); Chen, W.-L. [Department of Biomedical Imaging and Radiological Sciences, National Yang-Ming University, 155, Li-Nong Street Sec. 2, Pei-tou, Taipei 112, Taiwan (China)

    2008-02-15

    The thermoluminescent dosimeters (TLDs) are widely used for measuring the radiation dose. However, the luminescence centers of TLD disappeared by reading process, and then repetition of measurement is impossible. Radiophotoluminescent glass dosimeters (RPLGDs) can be repeatedly read and keep the luminescence centers for a long time. When machine errors occurred, RPLGD data can be re-analyzed to ensure the reliability of measurement results. Our previous study revealed that the RPLGD is one of the most important radiation dose measurement instruments as compared with TLD. Nevertheless, this RPLGD cannot measure the neutron dose. The aims of this study are to develop the novel compound of RPLGD to detect neutrons, and investigate their physical characteristics. In this study, some series self-fabricated glass dosimeters were prepared from reagent powders of AgCl, AgNO{sub 3}, AgPO{sub 3}, Al(OH){sub 3}, NaPO{sub 3}, Na{sub 2}CO{sub 3}, Na{sub 3}PO{sub 4} and P{sub 2}O{sub 5}. Based on this study, we found that the absorption spectra of irradiated glass wavelength maxima typically occurred in the 300-350 nm. Moreover, 0.1 mol% of sliver consist in our newly developed RPLGD showed the highest gamma ray detection sensitivity. The development of a novel compound RPLGD progress will be continuously improved in our laboratory.

  7. Research on the formula of radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Li Huazhi; Xiao Zhenhong; Lin Min; Cui Ying; Chen Kesheng; Chen Yundong; Ye Hongsheng; Lin Jingwen

    2006-10-01

    The formula of radiochromic film dosimeters was studied. Commercially available nylon was used as the matrix, while hexahydroxyethyl pararosaniline cyanide (HPR-CN) and pararosaniline cyanide (PR-CN) that are made in China and other countries were used as the dyes of the dosimeters. the performance of the thin film dosimeters made in CIAE was tested and compared with each other. The formula of the dosimeters was finally confirmed by testing its physical properties and dosimetric characteristics. (authors)

  8. Experimental ratio between the 'real' dose per organ and the calculated dose determined by means of the Embalse nuclear power plant's personal dosimeter

    International Nuclear Information System (INIS)

    Thomasz, E.; Salas, C.A.

    1987-01-01

    The specific purpose of the study was to determine the experimental ratio between the reading of dosimeters used by the personnel of the Embalse nuclear power plant and the 'real' dose absorbed by the worker in different organs. An anthropomorphic phantom ALDERSON internal and externally loaded with approximately 150 TLD crystals was used. This phantom was placed in five enclosures that were usually occupied by workers of the Embalse nuclear power plant. In this way, the average dose per organ and the effective equivalent dosis in each enclosure could be calculated and compared with the personal dosimeters placed over the thorax and the conversion factor rem/rem for each enclosure was determined. The average factor resulting from the five considered enclosures was 0.73 rem/rem. This means that the personal dosimeters over value the real dosis absorbed by the personnel of the Embalse nuclear power plant in approximately 37%. (Author)

  9. Experimental verification of a new neutron spectrometer for environmental dosimetry and area; Verficiacion experimental de un nuevo espectrometro de neutrones para dosimetria ambiental y de area

    Energy Technology Data Exchange (ETDEWEB)

    Gomez-Ros, J. M.; Romero, A.; Begogini, R.; Esposito, A.; Moraleda, M.; Lagares, J. I.; Sansaloni, F.; Arce, P.; Llop, J.

    2011-07-01

    In this communication, we present experimental results with a new neutron spectrometer, developed jointly by the Radiation Dosimetry Unit of CIEMAT Unita di Fisica and INFN-LNF Sanitary (Italy), consisting of a polyethylene moderating sphere detectors thermal neutrons (paired thermoluminescent dosimeters and activation foils) located in different positions. The device configuration and distribution of dosimeters are designed to elicit a response in a nearly isotropic up to 20 MeV energy range. (Author)

  10. Indoor radon level measurements in Iran using AEOI passive dosimeters

    International Nuclear Information System (INIS)

    Sohrabi, M.; Solaymanian, A.R.

    1988-01-01

    A passive radon diffusion dosimeter was developed at the RPD of AEOI for nationwide indoor radon level measurements. Several parameters of the dosimeter were studied. Radon levels were determined in about 250 houses in Ramsar (a high natural radiation area), Tehran, Babolsar and Gonabad. In this paper, the results of some dosimeter parameters as well as radon levels in indoor air are reported

  11. Calibration and testing of the DMG gamma dosimeter

    International Nuclear Information System (INIS)

    Dolgirev, E.I.

    1987-01-01

    25-1000 nGy/h (2.5-1000 μrad/h) absorbed dose gamma dosimeter for measuring the efficient equivalent irradiation dose for population is developed. It has two subranges 1000 nGy/h and 250 nGy/h. Results of dosimeter calibration and testing are presented. The dosimeter error for both subranges is less than 10%

  12. ATR neutron spectral characterization

    Energy Technology Data Exchange (ETDEWEB)

    Rogers, J.W.; Anderl, R.A.

    1995-11-01

    The Advanced Test Reactor (ATR) at INEL provides intense neutron fields for irradiation-effects testing of reactor material samples, for production of radionuclides used in industrial and medical applications, and for scientific research. Characterization of the neutron environments in the irradiation locations of the ATR has been done by means of neutronics calculations and by means of neutron dosimetry based on the use of neutron activation monitors that are placed in the various irradiation locations. The primary purpose of this report is to present the results of an extensive characterization of several ATR irradiation locations based on neutron dosimetry measurements and on least-squares-adjustment analyses that utilize both neutron dosimetry measurements and neutronics calculations. This report builds upon the previous publications, especially the reference 4 paper. Section 2 provides a brief description of the ATR and it tabulates neutron spectral information for typical irradiation locations, as derived from the more historical neutron dosimetry measurements. Relevant details that pertain to the multigroup neutron spectral characterization are covered in section 3. This discussion includes a presentation on the dosimeter irradiation and analyses and a development of the least-squares adjustment methodology, along with a summary of the results of these analyses. Spectrum-averaged cross sections for neutron monitoring and for displacement-damage prediction in Fe, Cr, and Ni are given in section 4. In addition, section4 includes estimates of damage generation rates for these materials in selected ATR irradiation locations. In section 5, the authors present a brief discussion of the most significant conclusions of this work and comment on its relevance to the present ATR core configuration. Finally, detailed numerical and graphical results for the spectrum-characterization analyses in each irradiation location are provided in the Appendix.

  13. Nuclear accident dosimeter designed for use with the Panasonic TLD system

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1985-01-01

    A new design for the nuclear accident dosimeter (NAD) compatible with the Panasonic TLD badge has recently been adopted for use at LLNL. This NAD was tested at the 1984 Oak Ridge National Laboratory Intercomparison of Criticality Accident Dosimeters study. We describe the procedures and constants developed to evaluate the NAD components. These constants were averaged to give reasonable results from bare and moderated spectra. Other procedures to evaluate a person's neutron dose using activation of the blood sodium and hair are described. These latter procedures are used to complement the dose determined using the NAD, or to determine a dose if a NAD had not been worn during exposure. If little is known about the configuration of the fissile material or shielding between the material and the exposed person, a procedure which combines the blood and hair activations gives a good estimate of the dose. (DT) 3 refs., 2 figs., 6 tabs

  14. Development and dosimetric evaluation of radiochromic PCDA vesicle gel dosimeters

    International Nuclear Information System (INIS)

    Sun, P.; Fu, Y.C.; Hu, J.; Hao, N.; Huang, W.; Jiang, B.

    2016-01-01

    The gel dosimeter has the unique capacity in recording radiation dose distribution in three dimensions (3D), which has the specific advantages in dosimetry measurements where steep dose gradients exist, such as in intensity-modulated radiation therapy (IMRT), brachytherapy and so on. Some 3D dosimeters, such as Fricke gel dosimeters, polymer gel dosimeters, the PRESAGE plastic dosimeters and micelle gel dosimeters have appeared recently. However, there are several disadvantages of these 3D dosimeters limit their application in radiotherapy dose verification. In this study, a novel radiochromic gel dosimeter for 3D dose verification of radiotherapy was developed by dispersing nanovesicles self-assembled by 10,12-pentacosadiynoic acid (PCDA) into the tissue equivalence gel matrix. The characteristics of radiochromic PCDA vesicle gel dosimeters were evaluated. The results indicate that these radiochromic gel dosimeters have good linear dose response to X-ray irradiation in the dose range of 2–100 Gy. In addition, the radiochromic gel dosimeters breakthrough the limitations of the existing gel dosimeters such as diffusion effect, post-radiation effect, and poor forming ability. The response of the gel dosimeter does not show any dose rate dependence, energy dependence and temperature effect, and there was no obvious difference in the gel response between single and cumulative dose of fractional irradiation. Hence, the radiochromic PCDA vesicle gel dosimeters developed in this study could be generally applied to 3D dose verification in radiotherapy. - Highlights: • A novel radiochromic gel dosimeter was developed by dispersing PCDA nanovesicles into the tissue equivalence gel matrix. • This nanovesicle overcomes the dose image blurring caused by the diffusion of monomer molecules. • This nanovesicle limits the polymer chain growth, so as to reduce the post-radiation effect. • The gel matrixes possess excellent tissue equivalence and elastic strength, which

  15. Investigating hydrogel dosimeter decomposition by chemical methods

    International Nuclear Information System (INIS)

    Jordan, Kevin

    2015-01-01

    The chemical oxidative decomposition of leucocrystal violet micelle hydrogel dosimeters was investigated using the reaction of ferrous ions with hydrogen peroxide or sodium bicarbonate with hydrogen peroxide. The second reaction is more effective at dye decomposition in gelatin hydrogels. Additional chemical analysis is required to determine the decomposition products

  16. Compton effect thermally activated depolarization dosimeter

    Science.gov (United States)

    Moran, Paul R.

    1978-01-01

    A dosimetry technique for high-energy gamma radiation or X-radiation employs the Compton effect in conjunction with radiation-induced thermally activated depolarization phenomena. A dielectric material is disposed between two electrodes which are electrically short circuited to produce a dosimeter which is then exposed to the gamma or X radiation. The gamma or X-radiation impinging on the dosimeter interacts with the dielectric material directly or with the metal composing the electrode to produce Compton electrons which are emitted preferentially in the direction in which the radiation was traveling. A portion of these electrons becomes trapped in the dielectric material, consequently inducing a stable electrical polarization in the dielectric material. Subsequent heating of the exposed dosimeter to the point of onset of ionic conductivity with the electrodes still shorted through an ammeter causes the dielectric material to depolarize, and the depolarization signal so emitted can be measured and is proportional to the dose of radiation received by the dosimeter.

  17. Clinical dosimeter based on diamond detector

    International Nuclear Information System (INIS)

    Chervjakov, A.M.; Ljalina, L.I.; Ljutina, G.J.; Khrunov, V.S.; Martynov, S.S.; Popov, S.A.

    2002-01-01

    Full text: Diamond detectors have found application in the relative dosimetry and their parameters have been described elsewhere. Today, the exclusive producer of the diamond detector is the Institute of Physical and Technical Problems, Russia, and exclusive dealer is the PTW-Freiburg. The main features of the diamond detector are good long time stability, suitable range of the energy dependence for photon and electron beams in clinical use, independence of the measured date from temperature and pressure. The high sensitivity per volume unit of the diamond detector (1500 times higher than ionization chamber) allowed using detectors with very small volume (1-5 mm 3 ) and rather simple electronics for ionization current registration. The new dosimeter consists of the diamond detector itself, 40 m registration cable, pre-amplifier, micro-processor block for data handling and absorbed dose calculation using the calibration factor of diamond detector in terms of absorbed dose to water. Dosimeter has the possibility to work with PC using standard RS-232 interface. The main features of the dosimeter are as follows: the range of dose rate measurements for photon, electron and proton beams is within 0.01-1.0 Gy/s; the energy ranges for photons are 0.08-25 MeV, and 4-25 MeV for electrons, with energy dependence no more than ±2%; the main uncertainty of the dose measurements is within ±2%; the pre-irradiation dose for diamond detector is no more than 10 Gy; the sensitive volume of the used diamond detectors is within 1-5 mm 3 ; the weight of the dosimeter no more than 2 kg. The new dosimeter was evaluated at the Central Research Institute of Roentgenology and Radiology, St. Petersburg, Russia to verify its performance. The dosimeter was used as a reference instrument for dose measurements at Cobalt-60 unit, SL75-5 and SL-20 linear accelerators and the test results have shown that the device have met the specifications. It is planned to produce dosimeter as serial device by

  18. Directional Radiation Dosimeter for Area and Environmental Monitoring

    International Nuclear Information System (INIS)

    Manzoli, J.E.; Campos, V.P.; Moura, E.S.

    2009-01-01

    It is presented a dosimeter that is able to measure the photon exposure and the direction from where the radiation came from. Preliminary measurements performed by this new directional radiation dosimeter demonstrate its application. This dosimeter consists of a small lead cube with thermoluminescent discs on each face, placed in well known coordinates. Only one dosimeter of this kind indicates the direction of the radiation beam, if it came from a unique position. This study was conducted inside the radiation room of a Cobalt-60 Gamma Irradiator and the dosimeter indicated the source position

  19. Dental tissue as a thermoluminescence dosimetry dosimeter

    International Nuclear Information System (INIS)

    Solaimani, F.; Zahmatkesh, M.H.; Akhlaghpoor, Sh.

    2003-01-01

    Background: Thermoluminescence dosimetry is one of the dosimetry procedures used widely as routine and personal dosimeters. In order to extend this kind of dosimeters, dental tissue has been examined and was found promising as a Thermoluminescence Dosimetry dosimeter. Materials and Methods: In this study, 70 health teeth were collected. The only criterion, wich was considered for selection of the teeth, was the healthiness of them regardless of age and gender of the donors. All collected samples were washed and cleaned and milled uniformly. The final powder had a uniform grain size between 100-300 micrometer. The sample was divided into four groups. Group A and B were used for measurement of density and investigation of variation of thermoluminescent characteristics with temperature respectively. Groups C and D were used for investigation of variation of thermoluminescent intensity with dose and fading of this intensity with time. In all cases the results obtained with dental tissue were compared to a standard LiF, thermoluminescence dosimetry dosimeter. Results: It was found that, average density of the dental tissue was 1.570 g/cm 3 , which is comparable to density of LiF, which is 1.612g/cm 3 . It was also concluded that the range of 0-300 d ig C , dental tissue has a simple curve with two specific peaks at 140 and 25 d ig C respectively. The experiment also showed that, the variation of relative intensity versus dose is linear in the range of 0.04-0.1 Gy. The fading rate of dental tissue is higher than LiF but still in the acceptable range (14% per month in compare to 5.2% per month). Conclusion: Dental tissue as a natural dosimeter is comparable with Thermoluminescence Dosimetry and can be used in accidental events with a good approximation

  20. Correlation of Neutron Data Taken at Commercial Nuclear Sites

    Energy Technology Data Exchange (ETDEWEB)

    Rathbun, L. A.; Endres, G. W.R.

    1983-10-01

    In this report, data from neutron measurement and dosimetry studies performed by Pacific Northwest Laboratory, the Environmental Measurements Laboratory, and Rensselaer Polytechnic Institute are examined and compared. The purpose of this data correlation effort is to determine whether useful relationships exist between the actual neutron dose equivalent in a typical commercial nuclear power reactor and various measurement parameters, such as ratios of the response of 9-in. to 3-in. spheres, neutron/gamma ratios, albedo dosimeter response and neutron spectrometer readings. In most neutron radiation fields found in the reactors visited, the response of albedo dosimeters can be brought into reasonable agreement with dose equivalents measured with multispheres, tissue equivalent proportional counters (TEPCs) or remmeters. Because the responses of the remmeters, like the responses of albedo dosimeters, are energy dependent, it is preferable to correct the responses of the albedo dosimeters to agree with dose equivalents measured with either TEPCs or multispheres. If one of these laboratory systems has been used to measure neutron dose equivalents at a specific pressurized water reactor, a calculated average albedo dosimeter correction factor can be used for most locations at that reactor. However, if the measured 9-in. to 3-in. remmeter ratio is greater than 0.20, it is advisable to use a plot of 9-in. to 3-in. remmeter ratios versus albedo dosimeter correction factors to obtain an albedo dosimeter correction factor. Because 9-in. to 3-in. remmeter ratios at boiling water reactors are typically greater than 0.20, the latter approach applies to this type of reactor.

  1. Different spectra with the same neutron source

    International Nuclear Information System (INIS)

    Vega C, H. R.; Ortiz R, J. M.; Hernandez D, V. M.; Martinez B, M. R.; Hernandez A, B.; Ortiz H, A. A.; Mercado, G. A.

    2010-01-01

    Using as source term the spectrum of a 239 Pu-Be source several neutron spectra have been calculated using Monte Carlo methods. The source term was located in the centre of spherical moderators made of light water, heavy water and polyethylene of different diameters. Also a 239 Pu-Be source was used to measure its neutron spectrum, bare and moderated by water. The neutron spectra were measured at 100 cm with a Bonner spheres spectrometer. Monte Carlo calculations were used to calculate the neutron spectra of bare and water-moderated spectra that were compared with those measured with the spectrometer. Resulting spectra are similar to those found in power plants with PWR, BWR and Candu nuclear reactors. Beside the spectra the dosimetric features were determined. Using moderators and a single neutron source can be produced neutron spectra alike those found in workplaces, this neutron fields can be utilized to calibrate neutron dosimeters and area monitors. (Author)

  2. The New HARSHAW Extremity Dosimeters for Gamma and Beta Ray Monitoring

    International Nuclear Information System (INIS)

    Fellinger, J.; Majewski, M.; Rotunda, J.; Tawi, R.

    1997-01-01

    Large personnel dosimetry services providing extremity monitoring with finger rings based on thermoluminescent detectors have long been looking for a practical method for automated reading including automated identification of the detectors.All existing methods are at least not very suitable for medical applications, particularly for surgery, due to the fact that cold sterilization is usually impossible.Bicron radiation Measurement Products developed in co-operation with the Austrian Research Centre Seibersdorf a new finger ring dosimeter DXT-RAD as a fast and economic solution for fully automated evaluation of extremity dosemeters. (authors)

  3. Review of the current deficiencies in personnel beta dosimetry, with recommendations

    International Nuclear Information System (INIS)

    Sherbini, S.; Porter, S.W.

    1983-06-01

    The report describes the design and use of personnel dosimeters used by the nuclear power industry to monitor occupational radiation exposure. It then shows why the monitoring method is inaccurate when personnel are exposed to beta particles. Then the report describes alternatives that would lead to improved measurements. The report also critiques the dosimetry processor testing criteria developed by the Health Physics Society Standards Committee. Survey instruments are shown often to be inaccurate when used to measure beta dose rates

  4. Personnel radiation dosimetry symposium: program and abstracts

    Energy Technology Data Exchange (ETDEWEB)

    1984-10-01

    The purpose was to provide applied and research dosimetrists with sufficient information to evaluate the status and direction of their programs relative to the latest guidelines and techniques. A technical program was presented concerning experience, requirements, and advances in gamma, beta, and neutron personnel dosimetry.

  5. Personnel radiation dosimetry symposium: program and abstracts

    International Nuclear Information System (INIS)

    1984-10-01

    The purpose was to provide applied and research dosimetrists with sufficient information to evaluate the status and direction of their programs relative to the latest guidelines and techniques. A technical program was presented concerning experience, requirements, and advances in gamma, beta, and neutron personnel dosimetry

  6. Quantitative relations between beta-gamma mixed-field dosimeter responses and dose-equivalent conversion factors according to the testing standard

    International Nuclear Information System (INIS)

    Gupta, V.P.

    1982-08-01

    The conventional two-element personnel dosimeters, usually having two thick TLD (thermoluminescent dosimetry) ribbons, are used extensively for radiation protection dosimetry. Many of these dosimeters are used for the measurement of beta and gamma radiation doses received in mixed beta-gamma fields. Severe limitations exist, however, on the relative magnitudes and energies of these fields that may be measured simultaneously. Moreover, due to a well-known energy dependence of these dosimeters, particularly for the beta-radiations, systematic errors will occur whenever the differences in workplaces and calibration radiation energies exist. A simple mathematical approach is presented to estimate the deep and shallow dose equivalent values at different energies for such dosimeters. The formulae correlate the dosimeter responses and dose equivalent conversion factors at different energies by taking into account the guidelines of the adopted ANSI Standard N13.11 and the dosimetry practices followed by most dosimeter processors. This standard is to be used in a mandatory testing program in the United States

  7. Development of a new personnel dosemeter with Ca So4:Dy+ PTFE

    International Nuclear Information System (INIS)

    Cardenete, C.M.

    1992-01-01

    A new personnel dosimeter was developed with substitution plane geometric in metallic filters. The evaluation method was based in linear comparison of responses over several filters. This method was tested, simulating fields of mixed energy, and it was obtained an average derivation of (11+-5)% and lower derivations to 20% in all calibration interval. (C.G.C.)

  8. CFRMF spectrum update and application to dosimeter cross-section data testing

    International Nuclear Information System (INIS)

    Anderl, R.A.; Harker, Y.D.; Millsap, D.A.; Rogers, J.W.; Ryskamp, J.M.

    1982-01-01

    The Coupled Fast Reactivity Measurements Facility (CFRMF) at the Idaho National Engineering Laboratory (INEL) is a Cross Section Evaluation Working Group (CSEWG) benchmark for data testing of dosimetry, fission-product and actinide cross sections important to fast-reactor technology. In this paper we present the results of our work in updating the CFRMF spectrum characterization and in applying CFRMF integral data to testing ENDF/B-V dosimeter cross sections. Updated characterization of the central neutron spectrum includes the results of neutronics calculations with ENDF/B-V nuclear data, the generation of a fine-group spectrum representation for integral data-testing applications, and a sensitivity and uncertainty analysis which provides a flux-spectrum covariance matrix related to uncertainties and correlations in the nuclear data used in a neutronics calculation. Our application of CFRMF integral data to cross section testing has included both conventional integral testing analyses and least-squares-adjustment analyses with the FERRET code. The conventional integral data-testing analysis, based on C/E ratios, indicates discrepancies outside the estimated integral test uncertainty for the 6 Li(n,He), 10 B(n,He), 47 Ti(n,p), 58 Fe(n,γ), 197 Au(n,γ) and 232 Th(n,γ) cross sections. The integral test uncertainty included contributions from the measured integral data and from the spectrum and cross sections used to obtain the calculated integral data. Within the uncertainty and correlation specifications for the input spectrum and dosimeter cross sections, the least-squares-adjustment analysis indicated a high degree of consistency between the measured integral data and the ENDF/B-V dosimeter cross sections for all reactions except 10 B

  9. Radiation cytogenetic in vitro studies on human donors in the development of a suitable biological dosimeter

    International Nuclear Information System (INIS)

    Barjaktarovic, N.

    1988-02-01

    The final report is on the work carried out under the Agency research contract 3173/RB entitled ''Radiation cytogenetic in vitro studies on human donors in the development of a suitable biological dosimeter'', at the Clinical Hospital Centre ''Zvezdara'' in Belgrade, Yugoslavia. In co-operation and co-ordination dissemination with an international team of cytogeneticists under the IAEA CRP, the development of a suitable biological dosimetry system has been accomplished at the national institute to assist reliably in the absorbed radiation-dose assessment of accidentally-over-exposed personnel. The quantitative yield of asymmetrical chromosomal aberrations, such as dicentrics, rings and fragments consequent to exposure(s) to radiation overdose, help in such estimation of vital prognostic and radiation protection significance. This biological dosimeter system is particularly essential where the exposed person was not wearing any physical dosemeter during the accident. Prerequisite for implementation of an effective biological dosimetry is the availability of a reliable standard dose-response curve and an adherence to a protocol for lymphocytic chromosome analysis in first division phase of lymphocytes. The validation of the reported biological dosimeter is established through its successful analysis of a simulated over-exposure incident, with the associated error of less than 10%. Analytical cytogenetic methods for whole- and part-body acute exposures have been discussed. Part of the results have been reported in the publications under the CRP concerned

  10. Radiation Dose Measurement Using Chemical Dosimeters

    International Nuclear Information System (INIS)

    Lee, Min Sun; Kim, Eun Hee; Kim, Yu Ri; Han, Bum Soo

    2010-01-01

    The radiation dose can be estimated in various ways. Dose estimates can be obtained by either experiment or theoretical analysis. In experiments, radiation impact is assessed by measuring any change caused by energy deposition to the exposed matter, in terms of energy state (physical change), chemical production (chemical change) or biological abnormality (biological change). The chemical dosimetry is based on the implication that the energy deposited to the matter can be inferred from the consequential change in chemical production. The chemical dosimetry usually works on the sample that is an aqueous solution, a biological matter, or an organic substance. In this study, we estimated absorbed doses by quantitating chemical changes in matter caused by radiation exposure. Two different chemical dosimeters, Fricke and ECB (Ethanol-Chlorobenzene) dosimeter, were compared in several features including efficacy as dose indicator and effective dose range

  11. Sensitive color dosimeters using photochromic diarylethenes

    International Nuclear Information System (INIS)

    Irie, Setsuko; Irie, Masahiro

    2008-01-01

    Various types of color dosimeters are conveniently used for estimating absorbed dose in the radiation sterilization of biomedical materials. Diarylethenes with heterocyclic aryl groups are extensively studied for the applications to the optoelectronic devices, such as optical memory media and photowitching devices because of their thermally irreversible and fatigue-resistant properties. The colors of diarylethenes never fade in the dark conditions. The thermally stable dithienylethene derivatives are applied to sensitive color dosimeters. Upon γ-irradiation, polystyrene films containing diarylethene derivatives, such as 1,2-bis(2-methyl-5-phenyl-3-thienyl) perfluorocyclopentene 1 or 1,2-bis(2,5-dimethyl-3-thienyl) perfluorocyclopentene 2, and fluorescent metal complexes turned blue or red. Even if the absorbed dose was as small as 10 Gy, a clear color change was observed. (author)

  12. Small is beautiful: SAIC's new dosimeter

    International Nuclear Information System (INIS)

    Benson, R.G.

    1991-01-01

    Science Applications International Corporation (California) has developed an energy-compensated Geiger tube in a package the size of a small pocket pager. In fact, the whole dosimeter measures just 48mm x 72mm x 17mm. The rugged, lightweight unit is sensitive enough to record radiation ranging from low background levels caused by the earth's surface, the sun, or cosmic radiation, to beyond lethal dose levels. The PD-1 provides dose measurement, dose rate measurement, and ''chip'' functions. A chirper sounds each time a specified dose is accumulated, and the chirp increments are defined by the user. A dosimeter reader provides a simple interface for bi-directional communication with host PC. The Geiger tube provides improved accuracy over a wider energy range than current solid state devices. Features such as long battery life, long calibration life (two years or longer), and easy calibration procedure should help to simplify the work of health physicists overseeing dosimetry management programmes. (author)

  13. Individual dosimeter for radon and thoron daughters

    International Nuclear Information System (INIS)

    Chapuis, A.M.; Duport, P.; Zettwoog, P.

    1979-01-01

    The dosimeter is designed for the continuous measurement of the concentration of α emitters from the uranium 238 and thorium 232 series. It enables the measurement of, firstly the aerosol concentration of 218 Po (Radium A), 214 Po (Radium C') and 212 Po (Thorium C') and secondly the activity of long-lived α emitters in aerosols coming from ore dusts. One light weight version of this dosimeter is autonomous for 18 hours and is designed to measure individual doses, due to inhalation, for workers employed in uranium mines and ore processing plants. An other version using the same sampling head allows the monitoring of air concentrations in working environments. Living quarters, or free air

  14. High dose potassium-nitrate chemical dosimeter

    International Nuclear Information System (INIS)

    Dorda de Cancio, E.M.; Munoz, S.S.

    1982-01-01

    This dosimeter is used to control 10 kGY-order doses (1 Mrad). Nitrate suffers a radiolitic reduction phenomena, which is related to the given dose. The method to use potassium nitrate as dosimeter is described, as well as effects of the temperature of irradiation, pH, nitrate concentration and post-irradiation stability. Nitrate powder was irradiated at a Semi-Industrial Plant, at Centro Atomico Ezeiza, and also in a Gammacell-220 irradiator. The dose rates used were 2,60 and 1,80 KGY/hour, and the given doses varied between 1,0 and 150 KGY. The uncertainty was +-3% in all the range. (author) [es

  15. Selfcalibrated alanine/EPR dosimeters. A new generation of solid state/EPR dosimeters

    International Nuclear Information System (INIS)

    Yordanov, N.D.; Gancheva, V.

    1999-01-01

    Alanine/EPR dosimeters are well established as secondary, reference dosimeters for high-energy radiation. However, there are various sources of uncertainty in the evaluation of absorbed dose. This arises primarily from the necessity to calibrate each EPR spectrometer and each batch of dosimeters before their use. In order to overcome this disadvantage, a new generation alanine/EPR dosimeter has been developed, and its possibilities as a radiation detector are reported. Principally, it is a mixture of alanine, some quantity of EPR active substance, and a binding material. The EPR active substance, acting as an internal EPR standard, is chosen to have EPR parameters which are independent of the irradiation dose. The simultaneous recording of the spectra of both the sample and the standard under the same experimental conditions and the estimation of the ratio I alanine /I Mn as a function of the absorbed dose strongly reduces the uncertainties. The response of these dosimeters for 60 Co γ-radiation exhibits excellent linearity and reproducibility in the range of absorbed dose, 10 2 - 5 x 10 4 Gy. (author)

  16. Phosphor for thermoluminescent type radiation dosimeter

    International Nuclear Information System (INIS)

    Nada, N.; Yamashita, T.

    1975-01-01

    This has the accumulation effect of radiation energy and is mainly used as the element for thermoluminescent type radiation dosimeters. It has as the principal constituent a phosphor consisting of calcium sulfate as the principal constituent and other impurity elements such as dysprosium, thulium and the like. It is more sensitive by the order of 1 to 2 or more figures than the conventional ones and is excellent in the retention of absorbed radiation energy. (U.S.)

  17. Dosimeter characteristics and service performance requirements

    International Nuclear Information System (INIS)

    Ambrosi, P.; Bartlett, D.T.

    1999-01-01

    The requirements for personal dosimeters and dosimetry services given by ICRP 26, ICRP 35, ICRP 60 and ICRP 75 are summarised and compared with the requirements given in relevant international standards. Most standards could be made more relevant to actual workplace conditions. In some standards, the required tests of energy and angular dependence of the response are not sufficient, or requirements on overall uncertainty are lacking. (author)

  18. Color-indicator dosimeter for ionizing radiation

    International Nuclear Information System (INIS)

    Panchenkov, G.M.; Kozlov, L.L.; Molin, A.A.; Ershova, Z.F.; Mikhailov, L.M.; Juzvyak, A.G.; Valitov, R.B.; Churov, V.P.; Grinev, M.P.

    1980-01-01

    Colorimetric dosimeter of ionizing radiation, containing 70-100 w % of a thermoplastic polymer, 10-40 w. % of a softener, 0.5-3.0 w. % of stabilizer and two dyes compatible with the polymer is designed. The first dye is chosen among zanthene- polymethine- or pyrazolon dyes, while the other is a triarylmethane- indigo- thiazine- indophenol- indiamine- or indaniline dye. (E.G.)

  19. Calibration of a TLD system to estimate personal doses in fields of gamma-neutrons radiation

    International Nuclear Information System (INIS)

    Villegas, E.N.; Somarriba, S.I.

    2016-01-01

    Currently Nicaragua has no personal neutron dosimetry system. The calibration of a batch of albedo neutron dosimeters consisting of two pairs of "6LiF and "7LiF (Mg, Ti) detectors was done. The dosimeter and reader sensitivities were obtained using a "1"3"7Cs source, and a neutron calibration factor was found with a "2"4"1AmBe source. Reproducibility and homogeneity tests were performed, and the detection limit of the system was determined. This calibration will allow the beginning of neutron personal monitoring in the country. (author)

  20. Conceptual design of the SMART dosimeter

    Science.gov (United States)

    Johnson, Erik B.; Vogel, Sam; Frank, Rebecca; Stoddard, Graham; Vera, Alonzo; Alexander, David; Christian, James

    2017-08-01

    Active dosimeters for astronauts and space weather monitors are critical tools for mitigating radiation induced health issues or system failure on capital equipment. Commercial spaceflight, deep space flight, and satellites require smarter, smaller, and lower power dosimeters. There are a number of instruments with flight heritage, yet as identified in NASA's roadmaps, these technologies do not lend themselves to a viable solution for active dosimetry for an astronaut, particularly for deep space missions. For future missions, nano- and micro-satellites will require compact instruments that will accurately assess the radiation hazard without consuming major resources on the spacecraft. RMD has developed the methods for growing an advanced scintillation material called phenylcarbazole, which provides pulse shape discrimination between protons and electrons. When used in combination with an anti-coincidence detector system, an assessment of the dose from charged ions and neutral particles can be determined. This is valuable as damage on a system (such as silicon or tissue) is dependent on the particle species. Using this crystal with readout electronics developed in partnership with COSMIAC at the University of New Mexico, the design of the Small Mixed field Autonomous Radiation Tracker (SMART) Dosimeter consists of a low-power analog to digital conversion scheme with low-power digital signal processing algorithms, which are to be implemented within a compact system on a chip, such as the Xilinx Zynq series. A review of the conceptual design is presented.

  1. MO-A-BRB-00: TG191: Clinical Use of Luminescent Dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2016-06-15

    This presentation will highlight the upcoming TG-191 report: Clinical Use of Luminescent Dosimeters. Luminescent dosimetry based on TLD and OSLD is a practical, accurate, and precise technique for point dosimetry in medical physics applications. The charges of Task Group 191 were to detail the methodologies for practical and optimal luminescent dosimetry in a clinical setting. This includes (1) To review the variety of TLD/OSL materials available, including features and limitations of each. (2) To outline the optimal steps to achieve accurate and precise dosimetry with luminescent detectors and to evaluate the uncertainty induced when less rigorous procedures are used. (3) To develop consensus guidelines on the optimal use of luminescent dosimeters for clinical practice. (4) To develop guidelines for special medically relevant uses of TLDs/OSLs (e.g., mixed field i.e. photon/neutron dosimetry, particle beam dosimetry, skin dosimetry). While this report provides general guidelines for arbitrary TLD and OSLD processes, the report, and therefore this presentation, provide specific guidance for TLD-100 (LiF:Ti,Mg) and nanoDot (Al2O3:C) dosimeters because of their prevalence in clinical practice. Learning Objectives: Understand the available dosimetry systems, and basic theory of their operation Understand the range of dose determination methodologies and the uncertainties associated with them Become familiar with special considerations for TLD/OSLD relevant for special clinical situations Learn recommended commissioning and QA procedures for these dosimetry systems.

  2. MO-A-BRB-01: TG191: Clinical Use of Luminescent Dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Kry, S. [UT MD Anderson Cancer Center (United States)

    2016-06-15

    This presentation will highlight the upcoming TG-191 report: Clinical Use of Luminescent Dosimeters. Luminescent dosimetry based on TLD and OSLD is a practical, accurate, and precise technique for point dosimetry in medical physics applications. The charges of Task Group 191 were to detail the methodologies for practical and optimal luminescent dosimetry in a clinical setting. This includes (1) To review the variety of TLD/OSL materials available, including features and limitations of each. (2) To outline the optimal steps to achieve accurate and precise dosimetry with luminescent detectors and to evaluate the uncertainty induced when less rigorous procedures are used. (3) To develop consensus guidelines on the optimal use of luminescent dosimeters for clinical practice. (4) To develop guidelines for special medically relevant uses of TLDs/OSLs (e.g., mixed field i.e. photon/neutron dosimetry, particle beam dosimetry, skin dosimetry). While this report provides general guidelines for arbitrary TLD and OSLD processes, the report, and therefore this presentation, provide specific guidance for TLD-100 (LiF:Ti,Mg) and nanoDot (Al2O3:C) dosimeters because of their prevalence in clinical practice. Learning Objectives: Understand the available dosimetry systems, and basic theory of their operation Understand the range of dose determination methodologies and the uncertainties associated with them Become familiar with special considerations for TLD/OSLD relevant for special clinical situations Learn recommended commissioning and QA procedures for these dosimetry systems.

  3. Study and characterization of dosimeter LiF:Mg,Cu,P for using in aeronautical dosimetry

    International Nuclear Information System (INIS)

    Flavia, Hanna; Federico, Claudio; Lelis, Odair; Pereira, Heloisa; Pereira, Marlon

    2014-01-01

    The effects of cosmic ionizing radiation incidents in aircraft components and crews has been a source of concern and motivated increasingly studies and improvements in the area. The low dose rates involved in this radiation field in aircraft flight altitudes imply Dosimetric necessity of using materials with high efficiency of detection, to enable studies lower cumulative doses resulting in shorter routes or lower altitude. The choice of thermoluminescent dosimeters LiF: Mg, Cu, P was done by having a detection efficiency of about fifteen times higher than its predecessor (LiF: Mg, Ti), and therefore, applied in very low doses dosimetry, and environmental dosimetry . The implementation of the use of pair dosimetric TLD-600H and 700H-TLD will serve as support for testing and studies on the effects of low doses of cosmic radiation in environmental dosimetry applied in the aviation environment in the usual flight altitudes. In this paper are presented the results of development of a methodology for dosimetry low doses of gamma radiation and neutrons using the pair dosimetric TLD-600H and 700H-TLD. The results demonstrate a sensitivity of dosimeters well above the dosimeters LiF: Mg, Ti confirming its suitability for dosimetry of low doses

  4. MO-A-BRB-01: TG191: Clinical Use of Luminescent Dosimeters

    International Nuclear Information System (INIS)

    Kry, S.

    2016-01-01

    This presentation will highlight the upcoming TG-191 report: Clinical Use of Luminescent Dosimeters. Luminescent dosimetry based on TLD and OSLD is a practical, accurate, and precise technique for point dosimetry in medical physics applications. The charges of Task Group 191 were to detail the methodologies for practical and optimal luminescent dosimetry in a clinical setting. This includes (1) To review the variety of TLD/OSL materials available, including features and limitations of each. (2) To outline the optimal steps to achieve accurate and precise dosimetry with luminescent detectors and to evaluate the uncertainty induced when less rigorous procedures are used. (3) To develop consensus guidelines on the optimal use of luminescent dosimeters for clinical practice. (4) To develop guidelines for special medically relevant uses of TLDs/OSLs (e.g., mixed field i.e. photon/neutron dosimetry, particle beam dosimetry, skin dosimetry). While this report provides general guidelines for arbitrary TLD and OSLD processes, the report, and therefore this presentation, provide specific guidance for TLD-100 (LiF:Ti,Mg) and nanoDot (Al2O3:C) dosimeters because of their prevalence in clinical practice. Learning Objectives: Understand the available dosimetry systems, and basic theory of their operation Understand the range of dose determination methodologies and the uncertainties associated with them Become familiar with special considerations for TLD/OSLD relevant for special clinical situations Learn recommended commissioning and QA procedures for these dosimetry systems.

  5. MO-A-BRB-00: TG191: Clinical Use of Luminescent Dosimeters

    International Nuclear Information System (INIS)

    2016-01-01

    This presentation will highlight the upcoming TG-191 report: Clinical Use of Luminescent Dosimeters. Luminescent dosimetry based on TLD and OSLD is a practical, accurate, and precise technique for point dosimetry in medical physics applications. The charges of Task Group 191 were to detail the methodologies for practical and optimal luminescent dosimetry in a clinical setting. This includes (1) To review the variety of TLD/OSL materials available, including features and limitations of each. (2) To outline the optimal steps to achieve accurate and precise dosimetry with luminescent detectors and to evaluate the uncertainty induced when less rigorous procedures are used. (3) To develop consensus guidelines on the optimal use of luminescent dosimeters for clinical practice. (4) To develop guidelines for special medically relevant uses of TLDs/OSLs (e.g., mixed field i.e. photon/neutron dosimetry, particle beam dosimetry, skin dosimetry). While this report provides general guidelines for arbitrary TLD and OSLD processes, the report, and therefore this presentation, provide specific guidance for TLD-100 (LiF:Ti,Mg) and nanoDot (Al2O3:C) dosimeters because of their prevalence in clinical practice. Learning Objectives: Understand the available dosimetry systems, and basic theory of their operation Understand the range of dose determination methodologies and the uncertainties associated with them Become familiar with special considerations for TLD/OSLD relevant for special clinical situations Learn recommended commissioning and QA procedures for these dosimetry systems.

  6. An ESR study on biological dosimeters: Human hair

    International Nuclear Information System (INIS)

    Colak, Seyda; Ozbey, Turan

    2011-01-01

    In the present work, characteristic features of the radicals found in untreated, gamma and UV-irradiated and mechanical damaged human hair samples were investigated by ESR spectroscopy. Heights of the resonance peaks measured with respect to the spectrum base line were used to monitor microwave power, dose-response, storage time and temperature dependent kinetic features of the radical species contributing to the formation of recorded experimental ESR spectra. Peak heights and g-values (2.0037-2.0052) determined from recorded spectra of hair were color dependent with ΔHpp-0.47 mT. The act of cutting hair samples gene rates sulfur centered radicals which are found in the a-keratin structure of hair. The variations of the peak heights with temperature were related with the water content found in the hair samples. In the 6-1100 Gy dose range, a linear + quadratic dose-response curve was recorded for hair and the mean radiation yield (G mean ) was calculated to be 0.4. The gamma radiation induced radicals were stable for a several hours at room temperature storage conditions. Based on these findings it was concluded that human hair samples could be used as biological/personnel dosimeters and that ESR spectroscopy could be successfully used as a potential technique for monitoring its dosimetric behaviours.

  7. Anthracene dosimeter characterization under radiotherapy photons

    International Nuclear Information System (INIS)

    Czelusniak, Caroline

    2011-01-01

    New radiotherapy techniques such as intensity-modulated radiation therapy and stereotactic radiosurgery have increased the need for dosimeters that can provide measurements in real time with high spatial resolution. Organic scintillation dosimeters are able to measure with accuracy small radiation fields and fields with high gradients, besides having advantages such as water and soft tissue equivalence and the possibility to be used in vivo. Anthracene is an organic scintillator crystal with the highest known scintillation efficiency among organic scintillation materials. The objective of this work is to characterize the anthracene as a dosimeter under radiotherapy photons energies, analysing its signal against average granulosity, intern capsule diameter, absorbed dose, absorbed dose rate, photon energy and its spatial resolution; with the last one analysed under three methods (edge spread function, line spread function and modulation transfer function). The photons energies used were 1.25 MeV ( 60 Co), 0.661 MeV ( 137 Cs) and X-rays (effective energies of 28.4; 46.5; 48.5; 94.0 e 106.0 keV). The scintillation detection system consisted of an optical fiber with one end attached to the anthracene capsule and the other to a photomultiplier tube maintained by power supply followed by an electrometer. Once Cerenkov radiation occurs in the optical fiber, it was removed from the total scintillation signal trough the subtraction of the signal, taken irradiating the optical fiber without the anthracene attached to one of its extremity. From results obtained, one can infer that the dosimeter signal increases proportionally with average granulosity and intern capsule diameter. The signal is linearly dependent of absorbed dose, linearly dependent of low photons energies and independent for high photons energies, as well as independent of the absorbed dose rate. From the spatial resolution values obtained it was possible to infer that the one obtained through modulation

  8. Comparison of electronic digital alarm dosimeter with TLD

    International Nuclear Information System (INIS)

    Kumar, Pankaj; Pandey, J.P.N.; Shinde, A M.; Purohit, R.G.; Sarkar, P.K.

    2012-01-01

    Control of exposure of radiation workers on day to day basis has been made easy by use of semiconductor based electronic digital dosimeter. Additional dose constraints of 10 mSv for occupational radiation workers have made it essential to use such type of digital personal monitoring devices. In addition to conventional ionisation chamber based direct reading dosimeters, additional 35 semiconductor based digital dosimeters model MGP DMC 2000 S were used for the monitoring of personal exposure of radiation workers in a spent fuel reprocessing plant. Though better least count and good performance over a wide range of dose rate are claimed by the manufacture, before making use of such dosimeter on large scale, validation of its performance is required to be checked. In this paper, an effort is made to determine the performance of digital dosimeters, by exposing these digital dosimeters in combination with TLDs at different radiation levels and obtained results were compared and analysed

  9. Criteria for personal dosimetry in mixed radiation fields in space. [analyzing trapped protons, tissue disintegration stars, and neutrons

    Science.gov (United States)

    Schaefer, H. J.

    1974-01-01

    The complexity of direct reading and passive dosimeters for monitoring radiation is studied to strike the right balance of compromise to simplify the monitoring procedure. Trapped protons, tissue disintegration stars, and neutrons are analyzed.

  10. Neutron stars

    International Nuclear Information System (INIS)

    Irvine, J.M.

    1978-01-01

    The subject is covered in chapters entitled: introduction (resume of stellar evolution, gross characteristics of neutron stars); pulsars (pulsar characteristics, pulsars as neutron stars); neutron star temperatures (neutron star cooling, superfluidity and superconductivity in neutron stars); the exterior of neutron stars (the magnetosphere, the neutron star 'atmosphere', pulses); neutron star structure; neutron star equations of state. (U.K.)

  11. Assessment of fast and thermal neutron ambient dose equivalents around the KFUPM neutron source storage area using nuclear track detectors

    Energy Technology Data Exchange (ETDEWEB)

    Fazal-ur-Rehman [Physics Department, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia)]. E-mail: fazalr@kfupm.edu.sa; Al-Jarallah, M.I. [Physics Department, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia); Abu-Jarad, F. [Radiation Protection Unit, Environmental Protection Department, Saudi Aramco, P. O. Box 13027, Dhahran 31311 (Saudi Arabia); Qureshi, M.A. [Center for Applied Physical Sciences, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia)

    2005-11-15

    A set of five {sup 241}Am-Be neutron sources are utilized in research and teaching at King Fahd University of Petroleum and Minerals (KFUPM). Three of these sources have an activity of 16Ci each and the other two are of 5Ci each. A well-shielded storage area was designed for these sources. The aim of the study is to check the effectiveness of shielding of the KFUPM neutron source storage area. Poly allyl diglycol carbonate (PADC) Nuclear track detectors (NTDs) based fast and thermal neutron area passive dosimeters have been utilized side by side for 33 days to assess accumulated low ambient dose equivalents of fast and thermal neutrons at 30 different locations around the source storage area and adjacent rooms. Fast neutron measurements have been carried out using bare NTDs, which register fast neutrons through recoils of protons, in the detector material. NTDs were mounted with lithium tetra borate (Li{sub 2}B{sub 4}O{sub 7}) converters on their surfaces for thermal neutron detection via B10(n,{alpha})Li6 and Li6(n,{alpha})H3 nuclear reactions. The calibration factors of NTD both for fast and thermal neutron area passive dosimeters were determined using thermoluminescent dosimeters (TLD) with and without a polyethylene moderator. The calibration factors for fast and thermal neutron area passive dosimeters were found to be 1.33 proton tracks cm{sup -2}{mu}Sv{sup -1} and 31.5 alpha tracks cm{sup -2}{mu}Sv{sup -1}, respectively. The results show variations of accumulated dose with the locations around the storage area. The fast neutron dose equivalents rates varied from as low as 182nSvh{sup -1} up to 10.4{mu}Svh{sup -1} whereas those for thermal neutron ranged from as low as 7nSvh{sup -1} up to 9.3{mu}Svh{sup -1}. The study indicates that the area passive neutron dosimeter was able to detect dose rates as low as 7 and 182nSvh{sup -1} from accumulated dose for thermal and fast neutrons, respectively, which were not possible to detect with the available active neutron

  12. Calibration of the polycarbonate dosimeter for the microdosimetry of 239Pu alpha particles in bone

    International Nuclear Information System (INIS)

    Stillwagon, G.B.; Morgan, K.Z.

    1977-01-01

    There has been some criticisms of the maximum permissible organ burden (MPOB) in bone for 239 Pu in recent years. These criticisms allude to the relative dearth of experimental data available concerning the actual dose delivered to the endosteal face of osseous tissue by the 239 Pu alpha particle. A dosimeter recently developed has been recommended for application to this microdosimetry problem. The tissue equivalence of polycarbonate dosimeters would allow dose equivalent to be read directly from the foil rather than determining activity from emulsions, in which the alpha particle range is different than in tissue, then relating this activity measurement to absorbed dose by some calculations. Although this dosimeter has been calibrated to read dose equivalent for fast neutron dosimetry, the need exists to determine the factor to multiply by the number of 239 Pu alpha-induced tracks to obtain dose equivalent. This problem is being approached in the following manner. A device called the vacuum-sealed alpha-calibrator has been designed and constructed which will allow the handling of a standard 239 Pu solution obtained for this purpose. The calibrator will first be connected to surface barrier detectors which feed data into a multi-channel analyzer. The counts obtained under the alpha peaks at various heights above the source and the accumulated time are input into a computer program recently written to convert this data into dose rate in rems/unit time. Next the measurements are duplicated, this time using the polycarbonate dosimeter. The results will produce a factor relating the number of alpha-induced tracks to dose

  13. Gadolinium oxide coated fully depleted silicon-on-insulator transistors for thermal neutron dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Vitale, Steven A., E-mail: steven.vitale@ll.mit.edu; Gouker, Pascale M.

    2013-09-01

    Fully depleted silicon-on-insulator transistors coated with gadolinium oxide are shown to be effective thermal neutron dosimeters. The theoretical neutron detection efficiency is calculated to be higher for Gd{sub 2}O{sub 3} than for other practical converter materials. Proof-of-concept dosimeter devices were fabricated and tested during thermal neutron irradiation. The transistor current changes linearly with neutron dose, consistent with increasing positive charge in the SOI buried oxide layer generated by ionization from high energy {sup 157}Gd(n,γ){sup 158}Gd conversion electrons. The measured neutron sensitivity is approximately 1/6 the maximum theoretical value, possibly due to electron–hole recombination or conversion electron loss in interconnect wiring above the transistors. -- Highlights: • A novel Gd{sub 2}O{sub 3} coated FDSOI MOSFET thermal neutron dosimeter is presented. • Dosimeter can detect charges generated from {sup 157}Gd(n,γ){sup 158}Gd conversion electrons. • Measured neutron sensitivity is comparable to that calculated theoretically. • Dosimeter requires zero power during operation, enabling new application areas.

  14. Blind test results - measure of incessant progress in personnel monitoring services at Kudankulam Nuclear Power Project

    International Nuclear Information System (INIS)

    Jashi, K.B.; Ashok Kumar, A.; Pandaram, P.

    2018-01-01

    Personnel Monitoring Services (PMS) at Thermoluminescent Dosimeter (TLD) laboratory, Kudankulam Nuclear Power Project (KKNPP) involve monitoring of about 1500 radiation workers of KKNPP for external radiation exposure on monthly basis. Quality assurance in PMS provides an overall estimate of accuracy which depends not only upon the dosimeter quality and reader performance but also on data interpretation, applicable software and the system. Therefore it measures the effectiveness of quality control mechanism. Blind test is an effective tool to check the performance of the processor without their knowledge during routine processing of the service cards

  15. IAEA reference dosimeter: Alanine-ESR

    International Nuclear Information System (INIS)

    Mehta, K.; Girzikowsky, R.

    1999-01-01

    Since 1985, the IAEA has been using alanine-ESR as a transfer dosimeter for its dose quality audit service, namely the International Dose Assurance Service. The alanine dosimeters are rod-type containing 70 wt% DL--α-alanine and 30 wt% polystyrene. We have two self-shielded gamma facilities for the calibration of the dosimetry system, where the temperature within the irradiation chamber can be controlled by a specially designed unit. A 4th order polynomial is fitted to the 16 data points in the dose range of 100 Gy to 50 kGy. The measured value of the irradiation temperature coefficient at two dose values (15 and 45 kGy) is 0.23 %/deg. C. Also, the ESR-response was followed for several dosimeters for about 8 months to study the post-irradiation effect. A value of 0.008 %/day was observed for the fading of the response for two dose values (15 and 45 kGy) and three irradiation temperatures (15, 27 and 40 deg. C). The effect of the analysis temperature on the ESR response was also studied. The combined relative uncertainty for the IAEA alanine-ESR dosimetry system is 1.5% (k=1). This includes that transferred from the primary laboratory for the dose rate measurements of the gamma facilities, dosimetry system calibration uncertainties, batch variability and uncertainty in the curve fitting procedure. This value however does not include the contribution due to the irradiation temperature correction which is applied when it differs from that during calibration; this component being specific for each dose measurement. (author)

  16. Acceptance Testing of Thermoluminescent Dosimeter Holders.

    Science.gov (United States)

    Romanyukha, Alexander; Grypp, Matthew D; Sharp, Thad J; DiRito, John N; Nelson, Martin E; Mavrogianis, Stanley T; Torres, Jeancarlo; Benevides, Luis A

    2018-05-01

    The U.S. Navy uses the Harshaw 8840/8841 dosimetric (DT-702/PD) system, which employs LiF:Mg,Cu,P thermoluminescent dosimeters (TLDs), developed and produced by Thermo Fisher Scientific (TFS). The dosimeter consists of four LiF:Mg,Cu,P elements, mounted in Teflon® on an aluminum card and placed in a plastic holder. The holder contains a unique filter for each chip made of copper, acrylonitrile butadiene styrene (ABS), Mylar®, and tin. For accredited dosimetry labs, the ISO/IEC 17025:2005(E) requires an acceptance procedure for all new equipment. The Naval Dosimetry Center (NDC) has developed and tested a new non-destructive procedure, which enables the verification and the evaluation of embedded filters in the holders. Testing is based on attenuation measurements of low-energy radiation transmitted through each filter in a representative sample group of holders to verify that the correct filter type and thickness are present. The measured response ratios are then compared with the expected response ratios. In addition, each element's measured response is compared to the mean response of the group. The test was designed and tested to identify significant nonconformities, such as missing copper or tin filters, double copper or double tin filters, or other nonconformities that may impact TLD response ratios. During the implementation of the developed procedure, testing revealed a holder with a double copper filter. To complete the evaluation, the impact of the nonconformities on proficiency testing was examined. The evaluation revealed failures in proficiency testing categories III and IV when these dosimeters were irradiated to high-energy betas.

  17. Evaluation of fading factor and self-dose for glass dosimeter and thermoluminescence dosimeter

    International Nuclear Information System (INIS)

    Yamasaki, T.; Yamanishi, H.; Miyake, H.; Komura, K.

    2000-01-01

    The glass dosimeter (GD) and thermoluminescence dosimeter (TLD) are both passive radiation detectors. They are often used for measuring environmental radiation. In order to measure low dose rate preciously, it is important to evaluate decreased dose due to fading and self-dose during the exposure period. We evaluate the fading factor and self-dose of thee passive detectors, GD and TLD. We select Ogoya tunnel for the experiment. The tunnel is suitable field for measuring faded dose and self-dose because it is low cosmic radiation. At the center of the tunnel, the intensity of cosmic ray is reduced to about 1/177 than the outside of the funnel. We prepared two sets of dosimeters. One set consists of five GDs, five TLDs and some pre-irradiated GDs and TLDs that are exposed to standard radiation of 4 mGy by Cs-137. These dosimeters are put in the 10 cm thick lead box in order to shield the terrestrial gamma ray. One set is located at the center of the tunnel and the other is the outside of the funnel. The dosimeters were exposed for ten months, from May 1998 to March 1999. After the exposure, the readers of dosimeters are carried into the funnel to read out the signals promptly as soon as taking out the dosimeters. As a result of the measurement, four kinds of data are taken for GD and TLD respectively. Assumed that the self-dose and cosmic ray are constant during exposure, the four independent unknown quantities, a self-dose a dose due to cosmic ray and a fading coefficient at the center of the tunnel and at the outside, are considered. Therefore four simultaneous equations should be obtained. From these examinations, the faded dose of GD is less than 1%, but that of TLD is about 16% during ten months. The coefficient for compensation of fading of GD and TLD is given as the half of the each value. At the outside of the tunnel, the measured dose rate of cosmic ray that can pass through the 10 cm lead is evaluated to be about 16 nGy/h by both detectors. The self

  18. Personnel Monitoring Department - DEMIN

    International Nuclear Information System (INIS)

    1989-01-01

    The activities and purposes of the Personnel Monitoring Dept. of the Institute of Radioprotection and Dosimetry of the Brazilian CNEN are presented. A summary of the personnel monitoring service is given, such as dosemeters supply, laboratorial inspections, and so on. The programs of working, publishing, courses and personnel interchange are also presented. (J.A.M.M.)

  19. Wallac automatic alarm dosimeter type RAD21

    International Nuclear Information System (INIS)

    Burgess, P. H.; Iles, W.J.

    1980-02-01

    The Automatic Alarm Dosimeter type RAD 21 is a batterypowered personal dosemeter and exposure rate alarm monitor, designed to be worn on the body, covering an exposure range from 0.1 to 999.9 mR and has an audible alarm which can be pre-set over the range 1 mR h -1 to 250 mR h -1 . The instrument is designed to measure x- and γ radiation over the energy range 50 keV to 3 MeV. The facilities and controls, the radiation, electrical, environmental and mechanical characteristics, and the manual, have been evaluated. (U.K.)

  20. Personnel preferences in personnel planning and scheduling

    OpenAIRE

    van der Veen, Egbert

    2013-01-01

    Summary The personnel of an organization often has two conflicting goals. Individual employees like to have a good work-life balance, by having personal preferences taken into account, whereas there is also the common goal to work efficiently. By applying techniques and methods from Operations Research, a subfield of applied mathematics, we show that operational efficiency can be achieved while taking personnel preferences into account. In the design of optimization methods, we explicitly con...

  1. Liquid polymers for using in a holographic ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Nicolau-Rebigan, S.

    1979-01-01

    Some liquid polymeric systems for using in the holographic ionizing radiation dosimeter are presented. It is shown that the action of radiation on polymers leads to the destruction of the polymeric chains or to perform them, the both processes being applied in radiation dosimetry. Some advantages of the holographic dosimeter are outlined comparatively with those common used. (author)

  2. Method of measuring neutron spectra in JMTR exclusively used for irradiation and their evaluation

    International Nuclear Information System (INIS)

    Sakurai, Kiyoshi

    1983-01-01

    In the core of the Japan Materials Testing Reactor, about 60 capsules are irradiated. These are the material capsules for irradiating reactor materials, the fuel capsules for irradiating reactor fuel, the RI capsules for producing radioisotopes and so on. In the irradiation experiment using a reactor, the information on the neutron fluence is indispensable, and the neutron fluence in the irradiated specimen part is evaluated with a dosimeter or the nuclear calculation for the core of the JMTR. At the time of irradiating reactor materials, the dosimeter Fe-54 (n,p) Mn-54 is generally used for evaluating the neutron fluence more than 1 MeV. In the case of fuel irradiation, the thermal neutron fluence is evaluated with the dosimeter Co-59 (n,γ) Co-60. It is important to examine in detail neutron spectra by both calculation and experiment in the reactors exclusively used for irradiation such as the JMTR. The neutron irradiation field in the JMTR, neutron spectrum measuring experiment, the neutron flux monitors for standardizing data, the measurement of X-ray and gamma ray, neutron guess spectrum, the compilation of neutron cross section for SAND 2, and the unfolding of neutron spectra are reported. The degree of agreement of the neutron fluence more than 1 MeV by measurement and calculation was +- 10 to 20 %. (Kako, I.)

  3. Assessment of Siemens plessey electronic personal dosimeter

    International Nuclear Information System (INIS)

    Hirning, C.R.; Lopez, S.; Yuen, P.S.

    1994-01-01

    This report presents the results of a laboratory assessment of the performance of a new type of personal dosimeter. The Electronic Personal Dosimeter, or EPD, was developed jointly by the National Radiological Protection Board and Siemens Plessey Controls Limited, both of the United Kingdom. Twenty pre-production units of the EPD and a reader were purchased by Ontario Hydro for the assessment. The tests were conducted jointly by Ontario Hydro's Health and Safety Division and AECL Research's Chalk River Laboratories (CRL), with funding from the Candu Owner's Group. A total of 26 tests were conducted, divided between Ontario Hydro and AECL. The test results were compared with the relevant requirements of three standards. In general, the performance of the EPD was found to be quite acceptable. It met most of the relevant requirements of the three standards and most of the design specifications. However, the following deficiencies were found: slow response time; sensitivity to high-frequency EMF; poor resistance to dropping; and an alarm that is not loud enough. In addition, the response of the EPD to low-energy beta rays may be too low for some applications. There were serious problems with the reliability of operation of the pre production EPDs used in these tests. 9 refs., 34 tabs., 20 figs

  4. Assessment of Siemens plessey electronic personal dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Hirning, C R; Lopez, S [Ontario Hydro, Toronto, ON (Canada); Yuen, P S [Atomic Energy of Canada Ltd., Chalk River, ON (Canada). Chalk River Nuclear Labs.

    1994-01-01

    This report presents the results of a laboratory assessment of the performance of a new type of personal dosimeter. The Electronic Personal Dosimeter, or EPD, was developed jointly by the National Radiological Protection Board and Siemens Plessey Controls Limited, both of the United Kingdom. Twenty pre-production units of the EPD and a reader were purchased by Ontario Hydro for the assessment. The tests were conducted jointly by Ontario Hydro`s Health and Safety Division and AECL Research`s Chalk River Laboratories (CRL), with funding from the Candu Owner`s Group. A total of 26 tests were conducted, divided between Ontario Hydro and AECL. The test results were compared with the relevant requirements of three standards. In general, the performance of the EPD was found to be quite acceptable. It met most of the relevant requirements of the three standards and most of the design specifications. However, the following deficiencies were found: slow response time; sensitivity to high-frequency EMF; poor resistance to dropping; and an alarm that is not loud enough. In addition, the response of the EPD to low-energy beta rays may be too low for some applications. There were serious problems with the reliability of operation of the pre production EPDs used in these tests. 9 refs., 34 tabs., 20 figs.

  5. The dosimeter personal use in controlled area

    International Nuclear Information System (INIS)

    Costa, R. F.

    2015-01-01

    The discovery of X-rays revolutionized medicine because it allowed a patient to be examined internally with no surgery. But also caused damage to health professionals and patients due, its oxidizing action. In the beginning of its discovery, many doctors were exposed and exposed beams to their patients for long periods of time, therefore, they developed diseases caused by radiation and the medical community realized that something was wrong. Then created a radiological protection commission to regulate its use in humans and so limit your exposure. Today we know that many companies still did not fit the standards of radiation protection. So we evaluate the technical professionals in radiology regarding the correct use of personal dosimeter, through a descriptive study with a quantitative approach, we used the information collection technique based on a questionnaire developed for this purpose which was delivered and collected personally. From this survey, we sought to assess the knowledge of the basic guidelines of radiological protection. He concluded that the majority of respondents know the rules of use of the personal dosimeter, but do not use it properly, due mainly to lack of supervision by the company, overwork and neglect. (author)

  6. Laser readable thermoluminescent radiation dosimeters and methods for producing thereof

    International Nuclear Information System (INIS)

    Braunlich, P.F.; Tetzlaff, W.

    1989-01-01

    Thin layer thermoluminescent radiation dosimeters for use in laser readable dosimetry systems, and methods of fabricating such thin layer dosimeters are disclosed. The thin layer thermoluminescent radiation dosimeters include a thin substrate made from glass or other inorganic materials capable of withstanding high temperatures and high heating rates. A thin layer of a thermoluminescent phosphor material is heat bonded to the substrate using an inorganic binder such as glass. The dosimeters can be mounted in frames and cases for ease in handling. Methods of the invention include mixing a suitable phosphor composition and binder, both being in particulate or granular form. The mixture is then deposited onto a substrate such as by using mask printing techniques. The dosimeters are thereafter heated to fuse and bond the binder and phosphor to the substrate. 34 figs

  7. Reactor Gamma Heat Measurements with Calorimeters and Thermoluminescence Dosimeters

    DEFF Research Database (Denmark)

    Haack, Karsten; Majborn, Benny

    1973-01-01

    Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than calorimet......Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than...... calorimeters, but possess advantages such as a small probe size and the possibility of making simultaneous measurements at many different positions. Hence, thermoluminescence dosimeters may constitute a valuable supplement to calorimeters for reactor γ-ray heating measurements....

  8. Energy dependence of fast neutron dosimetry using electrochemical etching

    International Nuclear Information System (INIS)

    Su, S.J.; Morgan, K.Z.

    1978-01-01

    Registration of fast-neutron induced recoil tracks by the electrochemical etching technique as applied to sensitive Lexan polycarbonate foils provides a simple and inexpensive means of fast neutron personnel dosimetry. The sensitivity (tracks/neutron) of recoil particle registration is given as a function of neutron energy. Neutrons of 7 Li (p,n) 7 Be, 3 T (d,n) 4 He and 9 B, respectively. Results are compared with other studies using other neutron sources and conventional etching method

  9. Temperature dependence of gafchromic MD-55 dosimeter

    International Nuclear Information System (INIS)

    Klassen, Norman V.; Zwan, Len van der; Cygler, Joanna

    1997-01-01

    Objective: Gafchromic MD-55 is a fairly new, thin film dosimeter that develops a blue color (λ max = 676 nm) when irradiated with ionizing radiation. The increase in absorbance is nearly proportional to the absorbed dose. MD-55 can be used for high precision dosimetry if care is taken to assure reproducible film orientation in the spectrophotometer as well as temperature control during both irradiation and reading. In order to achieve the maximum sensitivity of this dosimeter the readings of the optical density should be taken at λ max . It was reported for another type of Gafchromic film (DM-1260), that both λ max and ε max decrease with an increase in the temperature of the spectrophotometer. The purpose of this study was to characterize the reading temperature dependence of the new type of Gafchromic film available on the market and to find optimal conditions for using it for high precision dosimetry. Materials and Methods: Irradiations were carried out using 60 Co gamma rays from an Eldorado irradiator. The dosimeters were sandwiched in a lucite phantom with 4.4 mm build-up and irradiated in the center of a 10 cm x 10 cm field at 1 meter from the source. The temperature during irradiations was 22 deg. C. The dose rate was about 0.68 Gy/min. Measurements of optical density were made using a Cary 210 spectrophotometer. A bandpass of 3.5 nm was used. The temperature of the baseplate of the sample holder was regulated to +/-0.05 deg. C and measured by a probe lying on the baseplate. In all cases, values of OD were only recorded after they had come to a constant value, which was reached within 5 minutes of inserting the dosimeter into the sample chamber of the spectrophotometer. Results: The temperature dependence of the OD at 676 nm was measured in 2 studies using 6 dosimeters that had received 0, 1.0, 3.5, 6.2, 14.5 Gy. Readings were taken at 7 temperatures between 18.8 and 28.1 deg. C. By returning to the initial temperature several hours later, it was found

  10. Cytogenetic diagnostic of 3 populations of occupationally exposed personnel

    International Nuclear Information System (INIS)

    Guerrero C, C.; Arceo M, C.

    2013-10-01

    In the year 2000 the first service of biological dosimetry was requested to the Instituto Nacional de Investigaciones Nucleares (ININ), and until the year 2012 have been assisted 52 cases approximately. Most of the cases correspond to workers dedicated to the industrial radiography, followed by the occupationally exposed personnel either in the hospital area or health services and the minority corresponds to individuals linked to research institutions. The incident with more serious consequences to the individual happened to workers that ingested I-131 in the year 2003. Using the biological dosimetry to estimate exposure dose by damage in the lymphocyte chromosomes of each worker has been possible to establish the exposure dose in each one of them, or also to discard the supposed exposure. The dosimetry demonstrates to be an useful tool for situations with exposure suspicion, for example when the reading of thermoluminescent dosimeter of a occupationally exposed personnel does not correspond to the event, or when the personnel forgets to carry his dosimeter, the exposure dose can be determined. (Author)

  11. Fast neutron dosimetry by means of different solid state nuclear track detectors

    International Nuclear Information System (INIS)

    Spurny, F.; Turek, K.

    1977-01-01

    The comparative study of three different types of fast neutron dosimeters based on solid state nuclear track detectors is presented; the dosimeters studied were: - microscopic soda glass in contact with 232 Th; - polycarbonate Makrofol E; and - cellulose nitrate Kodak LR 115. All detectors were evaluated by visual counting in a microscope. The authors have studied such properties as the background, angular as well as energetical dependences of detectors. The results obtained show that all studied detectors are suitable for fast neutron dosimetry; their application depends however on the concrete experimental conditions (neutron spectrum, fluence etc.). Both advantages and disadvantages of each of them are presented. (Auth.)

  12. Luminescence studies of rare earth doped dosimeters

    International Nuclear Information System (INIS)

    Karali, T.

    1999-10-01

    The main objective of this thesis has been to address the applications and fundamentals of thermoluminescence (TL) and to contribute to existing knowledge about TL mechanisms in materials which are applied as radiation dosimeters. This issue has been explored for a long time but the mechanisms lack completeness and certainty. TL, Radioluminescence (RL) and Radio-thermoluminescence (RLTL) measurements have been conducted on a high sensitivity TL spectrometer both at low (30-290 K) and high (25-400 deg. C) temperatures, and different heat treatments (furnace and laser) were conducted in order to study the possible impurity clustering which changes the TL spectra and efficiency of the dosimeters. Studies have been based on three different host structure, namely sulphate, borates and zircon. The spectra of calcium sulphate samples doped with Tm 3+ and Dy 3+ at different concentration were examined using TL, RL and RLTL. Similar procedures were applied to the borate samples. Modifications of the material by thermal treatments convert the state of dispersion of the rare earth ions between isolated, pair or defect clusters, which alter the dosimeter efficiency. In some cases, modified geometries are detectable by movement of the line emissions such as for quenched samples which are attributable to new microcrystal line phases. The study of co-doped samples showed unequivocal evidence of a glow peak displacement of the two dopants within a single sample. This result supports the new view that RE 3+ ions could form part of a complex defect acting as both charge trap and recombination centres. Pulsed laser heating with a UV laser changed the glow curve shape and lead to strong signals. The detailed mechanisms for this process are discussed. The RL and TL spectra of synthetic zircon crystals doped with different RE 3+ ions (Pr, Sm, Eu, Gd, Ho, Dy, Er, and Yb) and phosphorus are reported. Even though there is some intrinsic emission from the host lattice the major signals are

  13. Measurement of epithermal neutrons by a coherent demodulation technique

    CERN Document Server

    Horiuchi, N; Takahashi, H; Kobayashi, H; Harasawa, S

    2000-01-01

    Epithermal neutrons have been measured using a neutron dosimeter via a coherent demodulation technique. This dosimeter consists of CsI(Tl)-photodiode scintillation detectors, four of which are coupled to neutron-gamma converting foils of various sizes. Neutron-gamma converting foils of In, Au and Co materials were used, each of which has a large capture cross section which peaks in the epithermal neutron energy region. The type of foil was selected according to the material properties that best correspond to the energy of the epithermal neutrons to be measured. In addition, the proposed technique was applied using Au-foils in order to measure the Cd ratio. The validity of the proposed technique was examined using an sup 2 sup 4 sup 1 Am-Be source placed in a testing stack of polyethylene blocks, and the results were compared with the theoretical values calculated by the Monte Carlo calculation. Finally, the dosimeter was applied for measuring epithermal neutrons and the Cd ratio in an experimental beam-tube o...

  14. Radiation doses to personnel in clinics for gynecologic oncology

    International Nuclear Information System (INIS)

    Forsberg, B.; Spanne, P.

    1985-01-01

    Radium or Cesium is used for radiotherapy of gynecologic cancer at six clinics in Sweden. This report gives a survey of the radiation doses the personnel is exposed to. The measurement were performed using TL-dosimeters. The dose equivalents for different parts of the body at specific working moments was deduced as well as the effective dose equivalent and the collective dose equivalent. 1983 the total collective dose equivalent for the six clinics was 1.3 manSv, which corresponds to 3.9 manmSv/g equivalent mass of Radium used at the treatments. (With 11 tables and 10 figures) (L.E.)

  15. Intercomparison of personal dose equivalent measurements by active personal dosimeters. Final report of a joint IAEA-EURADOS project

    International Nuclear Information System (INIS)

    2007-11-01

    Active personal dosimeters (APD) are widely used in many countries, i.e. in the medical field and as operational dosimeters in nuclear power plants. Their use as legal dosimeters is already established in a few countries, and will increase in the near future. In the majority of countries, APDs have not undergone accreditation programmes or intercomparisons. In 2001, an EURADOS (European Radiation Dosimetry Group) Working Group on harmonization of individual monitoring was formed, funded by the European Commission, in the fifth framework programme, and by the participating institutes. The work addressed four issues; inter alia also an inventory of new developments in individual monitoring with an emphasis on the possibilities and performance of active (electronic) dosimeters for both photon/beta and neutron dosimetry. Within the work on this issue, a catalogue of the most extensively used active personal dosimeters (APDs) suitable for individual monitoring was made. On the basis of the knowledge gained in this activity, the organization of an international intercomparison, which would address APDs, was considered of great value to the dosimetric community. The IAEA in cooperation with EURADOS organized such an intercomparison in which most of the testing criteria as described in two internationally accepted standards (IEC61526 and IEC61283) were used. Additionally, simulated workplace fields were used for testing the APD reactions to pulsed X ray fields and mixed gamma/X ray fields. This is the first time that results of comparisons of such types are published, which is of great importance for APD end users in medical diagnostic and surgery X ray applications. Nine suppliers from six countries in Europe and the USA participated in the intercomparison with 13 different models. One of the models was a special design for extremity dose measurements. Irradiations and readout was done by two accredited calibration laboratories in Belgium and France and the French

  16. An approved personal dosimetry service based on an electronic dosimeter

    International Nuclear Information System (INIS)

    Marshall, T.O.; Bartlett, D.T.; Burgess, P.H.; Campbell, J.I.; Hill, C.E.; Pook, E.A.; Sandford, D.J.

    1991-01-01

    At the Second Conference on Radiation Protection and Dosimetry a paper was presented which, in part, announced the development of an electronic dosimeter to be undertaken in the UK by the National Radiological Protection Board (NRPB) and Siemens Plessey Controls Ltd. This dosimeter was to be of a standard suitable for use as the basis of an approved personal dosimetry service for photon and beta radiations. The project has progressed extremely well and dosimeters and readers are about to become commercially available. The system and the specification of the dosimeter are presented. The NRPB is in the process of applying for approval by the Health and Safety Executive (HSE) to operate as personal monitoring service based on this dosimeter. As part of the approval procedure the dosimeter is being type tested and is also undergoing an HSE performance test and wearer trials. The tests and the wearer trials are described and a summary of the results to date presented. The way in which the service will be organized and operated is described and a comparison is made between the running of the service and others based on passive dosimeters at NRPB

  17. Light scattering in optical CT scanning of Presage dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Y; Adamovics, J; Cheeseborough, J C; Chao, K S; Wuu, C S, E-mail: yx2010@columbia.ed

    2010-11-01

    The intensity of the scattered light from the Presage dosimeters was measured using a Thorlabs PM100D optical power meter (Thorlabs Inc, Newton, NJ) with an optical sensor of 1 mm diameter sensitive area. Five Presage dosimeters were made as cylinders of 15.2 cm, 10 cm, 4 cm diameters and irradiated with 6 MV photons using a Varian Clinac 2100EX. Each dosimeter was put into the scanning tank of an OCTOPUS' optical CT scanner (MGS Research Inc, Madison, CT) filled with a refractive index matching liquid. A laser diode was positioned at one side of the water tank to generate a stationary laser beam of 0.8 mm width. On the other side of the tank, an in-house manufactured positioning system was used to move the optical sensor in the direction perpendicular to the outgoing laser beam from the dosimeters at an increment of 1 mm. The amount of scattered photons was found to be more than 1% of the primary light signal within 2 mm from the laser beam but decreases sharply with increasing off-axis distance. The intensity of the scattered light increases with increasing light attenuations and/or absorptions in the dosimeters. The scattered light at the same off-axis distance was weaker for dosimeters of larger diameters and for larger detector-to-dosimeter distances. Methods for minimizing the effect of the light scattering in different types of optical CT scanners are discussed.

  18. Water-resistant alanine-EPR dosimeter alanpol

    International Nuclear Information System (INIS)

    Peimel-Stuglik, Zofia; Bryl-Sandelewska, Teresa; Mirkowski, Krzysztof; Sartowska, Bozena

    2009-01-01

    Alanpol-water-resistant alanine-electron paramagnetic resonance (EPR) dosimeter consisted of cheap DL-α-alanine (9.8-27%) suspended in polyethylene matrix was presented. The rods (O=2.8 mm) were extruded from a hot mixture of alanine and low-density polyethylene. No grinding or crushing was used for alanine preparation. An orientation of cylindrical crystals, up to 300 μm long in parallel to the rod axis was responsible for some differences in a shape of EPR signal. These differences had no negative consequences for dosimetric applications. Signal-to-dose dependence was linear up to 10 kGy. Standard deviation of dosimetric answer was up to ±1.8% and up to 2.4% for dosimeters with 9.8% and 27% of DL-α-alanine, respectively. Irradiation temperature coefficient for both dosimeters was equal 0.2%/ deg. C. Hydrophobic properties of polyethylene and small number of alanine crystals located on the surface of the rod led to high resistance of dosimeters to water and humidity. The 24 h soaking of irradiated dosimeters in liquid water-reduced EPR signals by 3-4% and by 2-3% for dosimeters with 27% and 9.8% of DL-α-alanine, respectively. Three month storage time of irradiated dosimeters in room conditions decreases EPR signal for ∼3%.

  19. Investigating On Colour Stability Conditions Of Postirradiation Radiochromic Film Dosimeter

    International Nuclear Information System (INIS)

    Nguyen Nguyet Dieu; Doan Binh; Pham Thu Hong; Cao Van Chung; Nguyen Thanh Duoc

    2011-01-01

    B3 dosimeter is a thin film with average thickness of 0.0194 mm, which is supplied by the Gex company, the United States. This dosimeter was influenced by many factors: light, temperature, humidity during and after irradiation process. In fact, B3 film dosimeters will be stable under certain conditions such as tightly sealed packs, controlled irradiation and stored temperature after irradiated. Therefore, investigation of the stability effect of postirradiated B3 film dosimeters on the heating temperature, heating time and storing time is carried out before the absorbed dose is read and followed standard reading procedures. When exposed to ionizing radiation, the dosimeters change from colorless to colour. The absorbed doses are read on a Genesys 20 spectrophotometer at a wavelength of 544 nm. Absorbed dose range is investigated from 0.55 to 80 kGy. Experimental results were indicated that colour stability of the postirradiated dosimeters at a temperature of 65 ± 3 o C for 30 minutes and keeping them in desiccator for 5 minutes before read out. Under these conditions, colour stability of B3 film dosimeter has maintained for 3 months. (author)

  20. Use of Ukrainian semiconductor dosimeters in a CERN particle accelerator field

    Science.gov (United States)

    Rosenfeld, A.; Khivrich, V.; Kuts, V.; Tavlet, M.; Malfante, L.; Munoz-Ferrada, C.

    1994-08-01

    The results of the application of p-i-n and MOS dosimeters in the PS-ACOL Irradiation Facility (PSAIF) at CERN for separate measurements of gamma dose and fast neutron fluences are presented. The mixed gamma-neutron field was due to 26 GeV protons hitting an irridium target, yielding an instantaneous dose rate of approximately 3.10(sup 5) Gy/s. Good agreement with calibration curves is found for MOS in a (sup 60)Co gamma source as well as for p-i-n sensors in a neutron reactor spectrum with mean energy of 1 MeV. Experimental results from PSAIF are presented and pulse current injection annealing of p-i-n diodes is considered. Such sensors are very convenient for on-line separated total dose measurements in mixed gamma-neutron radiation fields, as well as for radiation hardness testing of electronic components on irradiation facilities, and could be installed near the detector area of LHC.

  1. New Generation of self-calibrated SS/EPR dosimeters: Alanine/EPR dosimeters

    International Nuclear Information System (INIS)

    Yordanov, N.D.; Gancheva, V.

    1999-01-01

    A new type of solid state/EPR dosimeters is described. Principally, it contains radiation sensitive diamagnetic material, some quantity of EPR active, but radiation insensitive, substance (for example Mn 2+ /MgO) and a binding material. In the present case alanine is used as a radiation sensitive substance. With this dosimeter, the EPR spectra of alanine and Mn 2+ are simultaneously recorded and the calibration graph represents the ratio of alanine versus Mn 2+ EPR signal intensity as a function of absorbed dose. In this way the reproducibility of the results is expected to be improved significantly including their intercomparison among different laboratories. Homogeneity of the prepared dosimeters and their behaviour (fading of EPR signals with time, influence of different meteorological conditions) show satisfactory reproducibility and stability with time. Because two different EPR active samples are recorded simultaneously, the influence of some instrument setting parameters (microwave power, modulation amplitude and modulation frequency) on the ratio I alanine /I Mn is also investigated. (author)

  2. p-MOSFET total dose dosimeter

    Science.gov (United States)

    Buehler, Martin G. (Inventor); Blaes, Brent R. (Inventor)

    1994-01-01

    A p-MOSFET total dose dosimeter where the gate voltage is proportional to the incident radiation dose. It is configured in an n-WELL of a p-BODY substrate. It is operated in the saturation region which is ensured by connecting the gate to the drain. The n-well is connected to zero bias. Current flow from source to drain, rather than from peripheral leakage, is ensured by configuring the device as an edgeless MOSFET where the source completely surrounds the drain. The drain junction is the only junction not connected to zero bias. The MOSFET is connected as part of the feedback loop of an operational amplifier. The operational amplifier holds the drain current fixed at a level which minimizes temperature dependence and also fixes the drain voltage. The sensitivity to radiation is made maximum by operating the MOSFET in the OFF state during radiation soak.

  3. SU-E-T-749: Thorough Calibration of MOSFET Dosimeters

    International Nuclear Information System (INIS)

    Plenkovich, D; Thomas, J

    2015-01-01

    Purpose: To improve the accuracy of the MOSFET calibration procedure by performing the measurement several times and calculating the average value of the calibration factor for various photon and electron energies. Methods: The output of three photon and six electron beams of Varian Trilogy linear accelerator SN 5878 was calibrated. Five reinforced standard sensitivity MOSFET dosimeters were placed in the calibration jig and connected to the Reader Module. As the backscatter material was used 7 cm of Virtual Water. The MOSFET dosimeters were covered with 1.5 cm thick bolus for the regular and SRS 6 MV beams, 3 cm bolus for 15 MV beam, 1.5 cm bolus for 6 MeV electron beam, and 2 cm bolus for the electron energies of 9, 12, 15, 18, and 22 MeV. The dosimeters were exposed to 100 MU, and the calibration factor was determined using the mobileMOSFET software. To improve the accuracy of calibration, this procedure was repeated ten times and the calibration factors were averaged. Results: As the number of calibrations was increasing the variability of calibration factors of different dosimeters was decreasing. After ten calibrations, the calibration factors for all five dosimeters were within 1% of one another for all energies, except 6 MV SRS photons and 6 MeV electrons, for which the variability was 2%. Conclusions: The described process results in calibration factors which are almost independent of modality or energy. Once calibrated, the dosimeters may be used for in-vivo dosimetry or for daily verification of the beam output. Measurement of the radiation dose under bolus and scatter to the eye are examples of frequent use of calibrated MOSFET dosimeters. The calibration factor determined for full build-up is used under these circumstances. To the best of our knowledge, such thorough procedure for calibrating MOSFET dosimeters has not been reported previously. Best Medical Canada provided MOSFET dosimeters for this project

  4. Organic liquids as ''activ media'' in a holographic ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Nicolau-Rebigan, S.

    1979-01-01

    Some types of organic liquids for using as activ media in a holographic ionizing radiation dosimeter are presented. One outlined the advantages of the holographic dosimeter comparatively with those of common used dosimeters. One presented the advantages of utilization of the organic liquids comparatively with another chemical systems used in a holographic ionizing radiation dosimeter. (author)

  5. Neutron measurement by transportable spectrometer

    International Nuclear Information System (INIS)

    Anon.

    1990-01-01

    Two levels of neutron spectrometry are in regular use at nuclear power plants: some techniques used in the laboratory produce detailed spectra but require specialist operators, while simple instruments used by non-specialists to measure the neutron dose-rate to operators provide little spectral information. The standard portable instruments are therefore of no use when anomalous readings are obtained which require further investigation. AEA Technology at Winfrith has developed a Transportable Neutron Spectrometer (TNS) which is designed to produce reasonable spectra in routine use by staff with no specialist skill in spectroscopy, and high-quality spectra in the hands of skilled staff. The TNS provides a level of information intermediate between those currently available, and is also designed to solve the problem of imperfect dose response which is common in portable dosimeters. The TNS system consists of a power supply, a probe and a signal processing and data acquisition unit. (author)

  6. Neutron dose and energy spectra measurements at Savannah River Plant

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Soldat, K.L.; Haggard, D.L.; Faust, L.G.; Tomeraasen, P.L.

    1987-08-01

    Because some workers have a high potential for significant neutron exposure, the Savannah River Plant (SRP) contracted with Pacific Northwest Laboratory (PNL) to verify the accuracy of neutron dosimetry at the plant. Energy spectrum and neutron dose measurements were made at the SRP calibrations laboratory and at several other locations. The energy spectra measurements were made using multisphere or Bonner sphere spectrometers, 3 He spectrometers, and NE-213 liquid scintillator spectrometers. Neutron dose equivalent determinations were made using these instruments and others specifically designed to determine dose equivalent, such as the tissue equivalent proportional counter (TEPC). Survey instruments, such as the Eberline PNR-4, and the thermoluminescent dosimeter (TLD)-albedo and track etch dosimeters (TEDs) were also used. The TEPC, subjectively judged to provide the most accurate estimation of true dose equivalent, was used as the reference for comparison with other devices. 29 refs., 43 figs., 13 tabs

  7. Development of a new type thyroid glands dosimeter

    International Nuclear Information System (INIS)

    He Lihua; Song Yiyang; Chen Qin; Chen Yannan

    2000-01-01

    A new dosimeter of 125 I in thyroid gland is described. The dosimeter consists of NaI(Tl) detector and intelligent data recorder. Single-chip-microcomputer is used for data handling. The activity of 125 I in thyroid glands of human being is measured directly, rapidly, and accurately. Furthermore, it can calculate and display the intake, committed dose equivalent and committed effective dose equivalent. The measuring range of 125 I in thyroid glands is 10-2 x 10 6 Bq. The dosimeter has been operating continuously for a long time with high stability

  8. Polymer gel dosimeter with AQUAJOINT® as hydrogel matrix

    Science.gov (United States)

    Maeyama, Takuya; Ishida, Yasuhiro; Kudo, Yoshihiro; Fukasaku, Kazuaki; Ishikawa, Kenichi L.; Fukunishi, Nobuhisa

    2018-05-01

    We report a polymer gel dosimeter based on a new gel matrix (AQUAJOINT®) that is a thermo-irreversible hydrogel formed by mixing two types of water-based liquids at room temperature. Normoxic N-vinylpyrrolidone-based polymer gels were prepared with AQUAJOINT® instead of gelatin. This AQUAJOINT®-based gel dosimeter exhibits a 2.5-fold increase in sensitivity over a gelatin-based gel dosimeter and a linear dose-response in the dose range of 0-8 Gy. This gel has heat resistance in a jar and controlled gel properties such as viscoelastic and mechanical characters, which may be useful for deformable polymer gel dosimetry.

  9. Environmental monitoring by CaSO4:Dy TL dosimeters

    International Nuclear Information System (INIS)

    Deme, S.; Szabo, P.P.

    1975-12-01

    The thermoluminescent dosimeters of high sensitivity are useful for monitoring the area near nuclear installations. CaSO 4 :Dy TL dosimeters have high sensitivity and low fading so that by means of them the dose from the background can be measured with an accuracy of 10-20%. An increase of 2 mR in the background can be observed and doses as high as 1000R can be registered with an accuracy of 5%. The measuring method and results are reported here. For two years these CaSO 4 :Dy dosimeters have been successfully used at the site of the Central Research Institute for Physics. (K.A.)

  10. Alanine EPR dosimeter response in proton therapy beams

    International Nuclear Information System (INIS)

    Gall, K.; Serago, C.; Desrosiers, M.; Bensen, D.

    1997-01-01

    We report a series of measurements directed to assess the suitability of alanine as a mailable dosimeter for dosimetry quality assurance of proton radiation therapy beams. These measurements include dose-response of alanine at 140 MeV, and comparison of response vs energy with a parallel plate ionization chamber. All irradiations were made at the Harvard Cyclotron Laboratory, and the dosimeters were read at NIST. The results encourage us that alanine could be expected to serve as a mailable dosimeter with systematic error due to differential energy response no greater than 3% when doses of 25 Gy are used. (Author)

  11. Investigation of self-indicating radiation personal dosimeter

    International Nuclear Information System (INIS)

    Xia Wen; Ye Honsheng; Lin Min; Xu Lijun; Chen Kesheng; Chen Yizhen

    2014-01-01

    A self-indicating radiation personal dosimeter was investigated using radiation sensitive material diacetylene monomer PCDA, which was a component of the polymerization system. The substrate material, solvent, sensitive material, solution temperature, thickness of film and the preparation method were studied. The dosimeter colour changes from white to blue when exposed 0.1-2.5 Gy, and the linearly dependent coefficient of the exposure response is 0.9998, the stability of absorbency in two weeks after exposure is testified well. It can be used as self-indicating radiation alert personal dosimeter. (authors)

  12. Device for the automatic evaluation of pencil dosimeters

    International Nuclear Information System (INIS)

    Schallopp, B.

    1976-01-01

    In connenction with the automation of radiation protection in nuclear power plants, an automatic reading device has been developed for the direct input of the readings of pencil dosimeters into a computer. Voltage measurements would be simple but are excluded, because the internal electrode of the dosimeter may not be touched, for operational reasons. This paper describes an optical/electronic conversion device in which the reading of the dosimeter is projected onto a Vidicon, scanned, and converted into a digital signal for output to the computer. (orig.) [de

  13. Personal noise dosimeters: Accuracy and reliability in varied settings

    Directory of Open Access Journals (Sweden)

    Sheri Lynn Cook-Cunningham

    2014-01-01

    Full Text Available This study investigated the accuracy, reliability, and characteristics of three brands of personal noise dosimeters (N = 7 units in both pink noise (PN environments and natural environments (NEs through the acquisition of decibel readings, Leq readings and noise doses. Acquisition periods included repeated PN conditions, choir room rehearsals and participant (N = 3 Leq and noise dosages procured during a day in the life of a music student. Among primary results: (a All dosimeters exhibited very strong positive correlations for PN measurements across all instruments; (b all dosimeters were within the recommended American National Standard Institute (ANSI SI.25-1991 standard of ±2 dB (A of a reference measurement; and (c all dosimeters were within the recommended ANSI SI.25-1991 standard of ±2 dB (A when compared with each other. Results were discussed in terms of using personal noise dosimeters within hearing conservation and research contexts and recommendations for future research. Personal noise dosimeters were studied within the contexts of PN environments and NEs (choral classroom and the day in the life of collegiate music students. This quantitative study was a non-experimental correlation design. Three brands of personal noise dosimeters (Cirrus doseBadge, Quest Edge Eg5 and Etymotic ER200D were tested in two environments, a PN setting and a natural setting. There were two conditions within each environment. In the PN environment condition one, each dosimeter was tested individually in comparison with two reference measuring devices (Ivie and Easera while PN was generated by a Whites Instrument PN Tube. In condition two, the PN procedures were replicated for longer periods while all dosimeters measured the sound levels simultaneously. In the NE condition one, all dosimeters were placed side by side on a music stand and recorded sound levels of choir rehearsals over a 7-h rehearsal period. In NE, condition two noise levels were measured

  14. Personnel preferences in personnel planning and scheduling

    NARCIS (Netherlands)

    van der Veen, Egbert

    2013-01-01

    Summary The personnel of an organization often has two conflicting goals. Individual employees like to have a good work-life balance, by having personal preferences taken into account, whereas there is also the common goal to work efficiently. By applying techniques and methods from Operations

  15. Estimation of the contribution of neutrons to the equivalent dose for personnel occupationally exposed and public in medical facilities: X-ray with energy equal or greater than 10MV

    International Nuclear Information System (INIS)

    Gonzalez, Alfonso Mayer; Jimenez, Roberto Ortega; Sanchez, Mario A. Reyes; Moranchel y Mejia, Mario

    2013-01-01

    In Mexico the use of electron accelerators for treating cancerous tumors had grown enormously in the last decade. When the treatments are carried out with X-ray beam energy below 10 MV the design of the shielding of the radioactive facility is determined by analyzing the interaction of X-rays, which have a direct impact and dispersion, with materials of the facility. However, when it makes use of X-ray beam energy equal to or greater than 10 MV the neutrons presence is imminent due to their generation by the interaction of the primary beam X-ray with materials head of the accelerator and of the table of treatment, mainly. In these cases, the design and calculation of shielding considers the generation of high-energy neutrons which contribute the equivalent dose that public and Occupationally Staff Exposed (POE) will receive in the areas surrounding the facility radioactive. However, very few measurements have been performed to determine the actual contribution to the neutron dose equivalent received by POE and public during working hours. This paper presents an estimate of the actual contribution of the neutron dose equivalent received by public and POE facilities in various radioactive medical use, considering many factors. To this end, measurements were made of the equivalent dose by using a neutron monitor in areas surrounding different radioactive installations (of Mexico) which used electron accelerators medical use during treatment with X-ray beam energy equal to or greater than 10 MV. The results are presented after a statistical analysis of a wide range of measures in order to estimate more reliability real contribution of the neutron dose equivalent for POE and the public. (author)

  16. Dosimetric properties of the pocket alarm dosimeter type Alnor RAD 21L, RAD 21H, RAD 22

    International Nuclear Information System (INIS)

    Hauser, M.; Burgkhardt, B.; Piesch, E.

    1981-02-01

    In personnel monitoring pocket dosimeters with build-in alarm devices are increasingly in use. The report presents results of a test performed at Karlsruhe for the pocket dose and alarm meter type Alnor RAD 21L, RAD 21H, RAD 22. The properties investigated are above all linearity and reproducibility of the dose reading as well as of the acoustic alarm indication, dependence of the dose reading on the photon energy, the direction of the radiation incidence, the dose rate, the temperature, operational characteristic of the batteries. (orig.) [de

  17. Personnel Policy and Profit

    DEFF Research Database (Denmark)

    Bingley, Paul; Westergård-Nielsen, Niels Chr.

    2004-01-01

    personnel structure variation. It is found that personnel policy is strongly related to economic performance. At the margin, more hires are associated with lower profit, and more separations with higher profit. For the average firm, one new job, all else equal, is associated with ?2680 (2000 prices) lower...

  18. Dose Determination using alanine detectors in a Mixed Neutron and Gamma Field for Boron Neutron Capture Therapy of Liver Malignancies

    DEFF Research Database (Denmark)

    Schmitz, T.; Blaickner, M.; Ziegner, M.

    2011-01-01

    Introduction Boron Neutron Capture Therapy for liver malignancies is being investigated at the University of Mainz. One important aim is the set-up of a reliable dosimetry system. Alanine dosimeters have previously been applied for dosimetry of mixed radiation fields in antiproton therapy, and ma...

  19. Facilities and procedures used for the performance testing of DOE personnel dosimetry systems

    Energy Technology Data Exchange (ETDEWEB)

    Roberson, P.L.; Fox, R.A.; Hogan, R.T.; Holbrook, K.L.; Hooker, C.D.; Yoder, R.C.

    1983-04-01

    Radiological calibration facilities for personnel dosimeter testing were developed at the Pacific Northwest Laboratory (PNL) for the Department of Energy (DOE) to provide a capability for evaluating the performance of DOE personnel dosimetry systems. This report includes the testing methodology used. The informational presented here meets requirements specified in draft ANSI N13.11 for the testing laboratory. The capabilities of these facilities include sealed source irradiations for /sup 137/Cs, several beta-particle emitters, /sup 252/Cf, and machine-generated x-ray beams. The x-ray beam capabilities include filtered techniques maintained by the National Bureau of Standards (NBS) and K-fluorescent techniques. The calibration techniques, dosimeter irradiation procedures, and dose-equivalent calculation methods follow techniques specified by draft ANSI N13.11 where appropriate.

  20. Facilities and procedures used for the performance testing of DOE personnel-dosimetry systems

    International Nuclear Information System (INIS)

    Roberson, P.L.; Fox, R.A.; Hogan, R.T.; Holbrook, K.L.; Hooker, C.D.; Yoder, R.C.

    1983-04-01

    Radiological calibration facilities for personnel dosimeter testing were developed at the Pacific Northwest Laboratory (PNL) for the Department of Energy (DOE) to provide a capability for evaluating the performance of DOE personnel dosimetry systems. This report includes the testing methodology used. The informational presented here meets requirements specified in draft ANSI N13.11 for the testing laboratory. The capabilities of these facilities include sealed source irradiations for 137 Cs, several beta-particle emitters, 252 Cf, and machine-generated x-ray beams. The x-ray beam capabilities include filtered techniques maintained by the National Bureau of Standards (NBS) and K-fluorescent techniques. The calibration techniques, dosimeter irradiation procedures, and dose-equivalent calculation methods follow techniques specified by draft ANSI N13.11 where appropriate

  1. A pocket warning γ-dosimeter with numerical display

    International Nuclear Information System (INIS)

    Jones, A.R.

    1980-09-01

    A pocket warning dosimeter is described. It provides alarms (continuous tone and a flashing red light) when a presettable dose has been accumulated in the range .064 - 16.4 rads (0.64 - 164 μGy). This warning level can be selected in nine steps of 2 with a switch inside the dosimeter. The dose rate is indicated by a series of sound pulses whose repetition rate is proportional to the dose rate. At 1 rad/h (10 mGy/h) about 17 pluses/minute are emitted. The accumulated dose up to 20 rads (0.2 Gy) is displayed in steps of 1 mrad (10 μGy) with a liquid crystal display. A red LED lights before battery failure occurs. The effects of changes in temperature, battery voltage, dose rate and photon energy upon dosimeter sensitivity are presented. Finally, the applications of the dosimeter are discussed. (auth)

  2. Investigating potential physicochemical errors in polymer gel dosimeters

    International Nuclear Information System (INIS)

    Sedaghat, Mahbod; Lepage, Martin; Bujold, Rachel

    2011-01-01

    Measurement errors in polymer gel dosimetry can originate either during irradiation or scanning. One concern related to the exothermic nature of polymerization reaction was that the heat released in polymer gel dosimeters during irradiation modifies their dose response. In this paper, the effect of heat released from the exothermal polymerization reaction on the dose response of a number of dosimeters was studied. In addition, we investigated whether heat-generated geometric distortion existed in newly proposed gel dosimeters that contain highly thermoresponsive polymers. Our results suggest that despite a significant internal temperature increase in some gel compositions, their dose responses are not affected when oxygen is well expelled mechanically from the gel mixture. We also report on significant pre-irradiation instability in some recently developed polymer gel dosimeters but that geometric distortions were not observed. Data obtained by a set of small calibration vials are compared to those obtained from larger phantoms, and potential physicochemical causes of deviations between them are identified.

  3. UVB DNA dosimeters analyzed by polymerase chain reactors

    International Nuclear Information System (INIS)

    Yoshida, Hiroko; Regan, J.D.; Florida Inst. of Tech., Melbourne, FL

    1997-01-01

    Purified bacteriophage λ DNA was dried on a UV-transparent polymer film and served as a UVB dosimeter for personal and ecological applications. Bacteriophage λ DNA was chosen because it is commercially available and inexpensive, and its entire sequence is known. Each dosimeter contained two sets of DNA sandwiched between UV-transparent polymer films, one exposed to solar radiation (experimental) and another protected from UV radiation by black paper (control). The DNA dosimeter was then analyzed by a polymerase chain reaction (PCR) that amplifies a 500 base pair specific region of λ DNA. Photoinduced damage in DNA blocks polymerase from synthesizing a new strand; therefore, the amount of amplified product in UV-exposed DNA was reduced from that found in control DNA. The dried λ DNA dosimeter is compact, robust, safe and transportable, stable over long storage times and provides the total UVB dose integrated over the exposure time. (author)

  4. Investigating potential physicochemical errors in polymer gel dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Sedaghat, Mahbod; Lepage, Martin [Centre d' imagerie moleculaire de Sherbrooke, Departement de medecine nucleaire et radiobiologie, Universite de Sherbrooke, Sherbrooke, QC (Canada); Bujold, Rachel, E-mail: martin.lepage@usherbrooke.ca [Service de radio-oncologie, Centre hospitalier universitaire de Sherbrooke, Sherbrooke, QC (Canada)

    2011-09-21

    Measurement errors in polymer gel dosimetry can originate either during irradiation or scanning. One concern related to the exothermic nature of polymerization reaction was that the heat released in polymer gel dosimeters during irradiation modifies their dose response. In this paper, the effect of heat released from the exothermal polymerization reaction on the dose response of a number of dosimeters was studied. In addition, we investigated whether heat-generated geometric distortion existed in newly proposed gel dosimeters that contain highly thermoresponsive polymers. Our results suggest that despite a significant internal temperature increase in some gel compositions, their dose responses are not affected when oxygen is well expelled mechanically from the gel mixture. We also report on significant pre-irradiation instability in some recently developed polymer gel dosimeters but that geometric distortions were not observed. Data obtained by a set of small calibration vials are compared to those obtained from larger phantoms, and potential physicochemical causes of deviations between them are identified.

  5. Temperature, humidity and time. Combined effects on radiochromic film dosimeters

    DEFF Research Database (Denmark)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 degrees C for irradiation by Co-60 photons and 10-Me......V electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is +0.......25 +/- 0.1% per degrees C for the FWT-60-00 dosimeters and +0.5 +/- 0.1% per degrees C For Riso B3 dosimeters at temperatures between 20 and 50 degrees C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger...

  6. Comparison of the effectiveness of polymer gel dosimeters (Magic ...

    African Journals Online (AJOL)

    demonstrate that the gel dosimeters are best suited for nuclear medicine. Keywords: Magic ... International Pharmaceutical Abstract, Chemical Abstracts, Embase, Index Copernicus, EBSCO, African. Index Medicus .... Reaction rate. 2.15E-6.

  7. Antioxidant effect of green tea on polymer gel dosimeter

    International Nuclear Information System (INIS)

    Samuel, E J J; Sathiyaraj, P; Deena, T; Kumar, D S

    2015-01-01

    Extract from Green Tea (GTE) acts as an antioxidant in acrylamide based polymer gel dosimeter. In this work, PAGAT gel was used for investigation of antioxidant effect of GTE.PAGAT was called PAGTEG (Polyacrylamide green tea extract gel dosimeter) after adding GTE. Free radicals in water cause pre polymerization of polymer gel before irradiation. Polyphenols from GTE are highly effective to absorb the free radicals in water. THPC is used as an antioxidant in polymer gel dosimeter but here we were replaced it by GTE and investigated its effect by spectrophotometer. GTE added PAGAT samples response was lower compared to THPC added sample. To increase the sensitivity of the PAGTEG, sugar was added. This study confirmed that THPC was a good antioxidant for polymer gel dosimeter. However, GTE also can be used as an antioxidant in polymer gel if use less quantity (GTE) and add sugar as sensitivity enhancer

  8. Storage Telemetry of Radionuclide Tracers by Implantable Thermoluminescent Dosimeters

    DEFF Research Database (Denmark)

    Bojsen, J.; Møller, U.; Christensen, Poul

    1977-01-01

    A storage telemetrical method using thermoluminescent (TL) dosimeters for long-term measurements of incorporated radioactive substances in unrestrained rats has been developed. The system has been used in combination with radiotelemetrical registration of the circadian temperature rhythm. By sequ...

  9. Effect of Gamma Rays on Fast Neutron Registration in CR-39

    CERN Document Server

    Kobzev, A P; El-Halem, A A; Abdul-Ghaphar, U S; Salama, T A

    2002-01-01

    A set of CR-39 plastic detectors with front PE radiator was exposed to Am-Be neutron source, which has an emission rate of 0.86\\cdot 10^{7} sec^{-1}, and the neutron dose equivalent rate 1 m apart from the source is equal to 11 mrem/hr. Another set of samples was irradiated by a neutron dose of 4 rem, then exposed to different gamma-ray doses using ^{60}Co source. It was found that the track density grows with the increase of neutron dose and etching time. It was also found that the bulk etching rate V_{B}, the track diameter and the sensitivity of the CR-39 plastic detector with respect to the neutron irradiation increased with increasing gamma-ray dose in the range 1?10 Mrad. These results show that CR-39 can be considered as a promising fast neutron dosimeter and gamma-ray dosimeter.

  10. Angular dependence of the nanoDot OSL dosimeter

    OpenAIRE

    Kerns, James R.; Kry, Stephen F.; Sahoo, Narayan; Followill, David S.; Ibbott, Geoffrey S.

    2011-01-01

    Purpose: Optically stimulated luminescent detectors (OSLDs) are quickly gaining popularity as passive dosimeters, with applications in medicine for linac output calibration verification, brachytherapy source verification, treatment plan quality assurance, and clinical dose measurements. With such wide applications, these dosimeters must be characterized for numerous factors affecting their response. The most abundant commercial OSLD is the InLight∕OSL system from Landauer, Inc. The purpose of...

  11. The NRPB's new dosimeter and dose record keeping services

    International Nuclear Information System (INIS)

    Dennis, J.A.; Marshall, T.O.; Shaw, K.B.

    1976-01-01

    A new automated dosimeter and record keeping service which the National Radiological Protection Board (UK) intends to introduce in 1977 is described. The automated system, based on a thermoluminescent dosimeter, will be linked to a fully computerised record keeping system with automatic printing of dose records and Transfer Records operated at its Headquarters at Harwell. The new system will dispense with much manual labour which in the past has introduced inevitable errors and incurred increasing costs. (U.K.)

  12. Polarized neutrons

    International Nuclear Information System (INIS)

    Williams, W.G.

    1988-01-01

    The book on 'polarized neutrons' is intended to inform researchers in condensed matter physics and chemistry of the diversity of scientific problems that can be investigated using polarized neutron beams. The contents include chapters on:- neutron polarizers and instrumentation, polarized neutron scattering, neutron polarization analysis experiments and precessing neutron polarization. (U.K.)

  13. Calibration and performance testing of electronic personal dosimeters (EPD)

    International Nuclear Information System (INIS)

    Banaga, H.A.

    2008-04-01

    In modern radiation protection practices, active personal dosimeters are becoming absolutely necessary operational tools for satisfying the ALARA principle. The aim of this work was to carry out calibration and performance testing of ten electronic personal dosimeters (EPD) used for the individual monitoring. The EPDs were calibrated in terms of operation radiation protection quantity, personal dose equivalent, Hp (10). Calibrations were carried out at three of x-ray beam qualities described in ISO 4037 namely 60, 100 and 150 kV in addition to Cs-137 gamma ray quality. The calibrations were performed using polymethylmethacrylate (PMMA) phantom with dimensions 20*20*15 cm 3 . Conversion coefficient Hp (10)/K air for the phantom was also calculated. The response and linearity of the dosimeter at the specified energies were also tested. The EPDs tested showed that the calibration coefficient ranged from 0.60 to 1.31 and an equivalent response for the specified energies that ranged from 0.76 to 1.67. The study demonstrated the possibility of using non standard phantom for calibrating dosimeters used for individual monitoring. The dosimeters under study showed a good response in all energies except the response in quality 100 kV. The linearity of the dosimeters was within ±15%, with the exception of the quality 100 kV where this limit was exceeded.(Author)

  14. Characterization of Thymol blue Radiochromic dosimeters for high dose applications

    Directory of Open Access Journals (Sweden)

    Feras M. Aldweri

    2018-03-01

    Full Text Available Thymol blue (TB solutions and Thymol blue Polyvinyl Alcohol (TB-PVA films have been introduced as Radiochromic dosimeter for high dose applications. The dosimeters were irradiated with gamma ray (60Co source from 5 to 30 kGy for film, and from 0.150 kGy to 4 kGy for solution. The optical density of unirradiated and irradiated TB solution as well as TB-PVA film dosimeters were studied in terms of absorbance at 434 nm using UV/VIS spectrophotometer. The effects of scan temperature, light pre-gamma irradiation, dose rate, relative humidity and stability of the absorbance of solutions and films after irradiation were investigated. We found the dose sensitivity of TB solution and TB-PVA film dosimeters increases significantly with increases of the absorbed dose as well as with the increases of TB dye concentrations. The useful dose range of developed TB solutions and TB-PVA films dosimeters is in the range 0.125–1 kGy and of 5–20 kGy, respectively. Keywords: Dose sensitivity, Radio-chromic dosimeter, Thymol blue, Absorbance, Concentrations

  15. Experimental evaluation of a MOSFET dosimeter for proton dose measurements

    International Nuclear Information System (INIS)

    Kohno, Ryosuke; Nishio, Teiji; Miyagishi, Tomoko; Hirano, Eriko; Hotta, Kenji; Kawashima, Mitsuhiko; Ogino, Takashi

    2006-01-01

    The metal oxide semiconductor field-effect transistor (MOSFET) dosimeter has been widely studied for use as a dosimeter for patient dose verification. The major advantage of this detector is its size, which acts as a point dosimeter, and also its ease of use. The commercially available TN502RD MOSFET dosimeter manufactured by Thomson and Nielsen has never been used for proton dosimetry. Therefore we used the MOSFET dosimeter for the first time in proton dose measurements. In this study, the MOSFET dosimeter was irradiated with 190 MeV therapeutic proton beams. We experimentally evaluated dose reproducibility, linearity, fading effect, beam intensity dependence and angular dependence for the proton beam. Furthermore, the Bragg curve and spread-out Bragg peak were also measured and the linear-energy transfer (LET) dependence of the MOSFET response was investigated. Many characteristics of the MOSFET response for proton beams were the same as those for photon beams reported in previous papers. However, the angular MOSFET responses at 45, 90, 135, 225, 270 and 315 degrees for proton beams were over-responses of about 15%, and moreover the MOSFET response depended strongly on the LET of the proton beam. This study showed that the angular dependence and LET dependence of the MOSFET response must be considered very carefully for quantitative proton dose evaluations

  16. Radiation measured for ISS-Expedition 12 with different dosimeters

    International Nuclear Information System (INIS)

    Zhou, D.; Semones, E.; Gaza, R.; Johnson, S.; Zapp, N.; Weyland, M.

    2007-01-01

    Radiation in low Earth orbit (LEO) is mainly from Galactic Cosmic Rays (GCR), solar energetic particles and particles in South Atlantic Anomaly (SAA). These particles' radiation impact to astronauts depends strongly on the particles' linear energy transfer (LET) and is dominated by high LET radiation. It is important to investigate the LET spectrum for the radiation field and the influence of radiation on astronauts. At present, the best active dosimeters used for all LET are the tissue equivalent proportional counter (TEPC) and silicon detectors; the best passive dosimeters are thermoluminescence dosimeters (TLDs) or optically stimulated luminescence dosimeters (OSLDs) for low LET and CR-39 plastic nuclear track detectors (PNTDs) for high LET. TEPC, CR-39 PNTDs, TLDs and OSLDs were used to investigate the radiation for space mission Expedition 12 (ISS-11S) in LEO. LET spectra and radiation quantities (fluence, absorbed dose, dose equivalent and quality factor) were measured for the mission with these different dosimeters. This paper introduces the operation principles for these dosimeters, describes the method to combine the results measured by CR-39 PNTDs and TLDs/OSLDs, presents the experimental LET spectra and the radiation quantities

  17. Neutron--neutron logging

    International Nuclear Information System (INIS)

    Allen, L.S.

    1977-01-01

    A borehole logging tool includes a steady-state source of fast neutrons, two epithermal neutron detectors, and two thermal neutron detectors. A count rate meter is connected to each neutron detector. A first ratio detector provides an indication of the porosity of the formation surrounding the borehole by determining the ratio of the outputs of the two count rate meters connected to the two epithermal neutron detectors. A second ratio detector provides an indication of both porosity and macroscopic absorption cross section of the formation surrounding the borehole by determining the ratio of the outputs of the two count rate meters connected to the two thermal neutron detectors. By comparing the signals of the two ratio detectors, oil bearing zones and salt water bearing zones within the formation being logged can be distinguished and the amount of oil saturation can be determined. 6 claims, 2 figures

  18. Perofrmance testing of personnel dosimetry services. Final report of a two-year pilot study, October 1977-September 1979

    International Nuclear Information System (INIS)

    Plato, P.; Hudson, G.

    1980-01-01

    A two-year pilot study was conducted of the Health Physics Society Standards Committee (HPSSC) Standard titled, Criteria for Testing Personnel Dosimetry Performance. The objectives of the pilot study were: to give processors an opportunity to correct any problems that are uncovered; to develop operational and administrative prodedures to be used later by a permanent testing laboratory; and to determine whether the proposed HPSSC Standard provides an adequate and practical test of dosimetry performance. Fifty-nine dosimetry processors volunteered to submit dosimeters for test irradiations according to the requirements of the HPSSC Standard. The feasibility of using the HPSSC Standard for a future mandatory testing program for personnel dosimetry processors is discussed. This report shows the results of the pilot study and contains recommendations for revisions in the Standard that will make a mandatory testing program useful to regulatory agencies, dosimetry processors, and radiation workers that use personnel dosimeters

  19. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR; Automatizacion de la calibracion de dosimetros de radioterapia en el laboratorio secundario de calibracion dosimetrica del CPHR

    Energy Technology Data Exchange (ETDEWEB)

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez, E-mail: c19btm@frcuba.co.cu [Centro de Proteccion e Higiene de las Radiaciones (CPHR), La Habana (Cuba)

    2013-11-01

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba.

  20. Training of nonlicensed personnel

    International Nuclear Information System (INIS)

    Hetrick, D.E.

    1975-01-01

    The safety and efficiency with which a station operates is a function of the competence and proficiency of all personnel. This includes the nonlicensed personnel who make up the bulk of the station staff. Thus the training of these members of the station complement is an important function in overall station performance. Standards, regulations, regulatory guides, and codes provide guidance to the training requirements for such personnel. Training needs and objectives must be established, a plan prepared and then all incorporated into a training program. A well planned and operated training program will stimulate effective communications between the different groups within the station and between the station and off site support groups

  1. Development of a multichannel dosimeter for radiotherapy

    International Nuclear Information System (INIS)

    Menezes, Claudio Jose Mesquita

    2000-06-01

    In radiotherapy, verification of the patient dose is of great important for the success of the treatment. Uncertainties in the evaluation of this dose can produce serious complications such as the loss of the control of the disease and damage to normal tissue. Semiconductor detectors present advantages over other types of radiation detectors such as ionization chambers and thermoluminescent dosimeters including small dimensions, high sensitivity and fast response. In this work, a multichannel dosimetric system is linear with dose, for a 6 MV x-ray beam and also with a beam of cobalt-60 gamma rays. The coefficients of determination of the calibration curves were better then 0,9998 in all cases. The four sensors presented similar response with the dose for different field sizes. The variation of the response was smaller than 1%. In a related study, depth dose was measured, and the results showed a good agreement compared to theoretical values. The angular response of the detectors showed a variation of 7% for angles of 45 deg C. Using the Anderson Random phantom, dose at the isocenter was determined from measurements of the surface dose. From the results obtained it can be concluded that the dosimetric system developed is adequate for the evaluation of many parameters in radiation fields used in radiotherapy. This system can be used to measure the patient entrance dose under treatment conditions, and the equipment can be used in the radiotherapy quality assurance program. (author)

  2. Commissioning and characteristics of MOSFET dosimeter

    International Nuclear Information System (INIS)

    Gopiraj, A.; Billimagga, Ramesh S.; Rekha, M.; Ramasubramaniam, V.

    2007-01-01

    The verification of the dose delivered to a patient is an important part of the quality assurance in radiotherapy. Thermoluminescent dosimeters (TLDs) and semiconductor diodes were mostly used for this purpose. Recently Metal Oxide Semiconductor field effect transistors (MOSFET) have been proposed for the application in radiotherapy. Each type of detector has its own advantages and disadvantages. The TLD size is very small and therefore can be used both for measurement and dose delivered to a patient and for measurements of dose distribution in a humanoid phantom. The main disadvantages of the TLDs are the time required by the preparation procedure and the limited accuracy which depends on the experience of the user. Additionally, TLDs do not allow an immediate readout. The main disadvantages of semiconductor diodes are the necessity of using a cable which can disturb normal clinical work especially when in vivo measurements are carried out, and the necessity of applying of many correction factors to achieve high accuracy. We procured MOSFET system from Thomson and Nielsen Electronic Ltd. The reproducibility as a function of dose and linearity and calibration factor of the MOSFET detectors were measured. The effects of energy, field size and accumulated dose on the response of the detectors were investigated

  3. Excited species in the FBX dosimeter system

    International Nuclear Information System (INIS)

    Gupta, B.L.

    2003-01-01

    In the FBX dosimeter solution, the excitation of xylenol orange (XO) produces maximum emission at 550-575 nm both at room and liquid nitrogen temperatures (about 85%) having a lifetime of 0.20-0.36 ns. In addition, at room temperature there is an emission at 350 nm for the excitation at 260 nm (about 15%) having a longer lifetime of 3.71-4.01 ns. Benzoic acid (BA) has excitation at 284-295 nm and emission at 320-365 nm having a lifetime of 1.38 ns. In an aqueous solution containing 5x10 -3 mol dm -3 BA, 2x10 -4 mol dm -3 XO and 0.04 mol dm -3 H 2 SO 4 there is no XO emission at 550 nm due to UV absorption at 260 nm by BA. In this solution, 2 emissions are observed near 350-360 nm, having lifetimes of 1.25 ns (89%) and 2.86 ns (11%). The wavelengths for the emission of XO and absorption of ferric-XO complex are nearly the same. Excited XO produces oxidation of ferrous ions and BA increases the chain length

  4. Evaluation of environmental monitoring thermoluminescent dosimeter locations

    International Nuclear Information System (INIS)

    Kinnison, R.

    1992-12-01

    Geostatistics, particularly kriging, has been used to assess the adequacy of the existing NTS thermoluminescent dosimeter network for determination of environmental exposure levels. (Kriging is a linear estimation method that results in contour plots of both the pattern of the estimated gamma radiation over the area of measurements and also of the standard deviations of the estimated exposure levels.) Even though the network was not designed as an environmental monitoring network, ft adequately serves this function in the region of Pahute and Rainier Mesas.. The Yucca Flat network is adequate only if a reasonable definition of environmental exposure levels is required; R is not adequate for environmental monitoring in Yucca Flat if a coefficient of variation of 10 percent or less is chosen as the criterion for network design. A revision of the Yucca Flat network design should be based on a square grid pattern with nodes 5000 feet (about one mile) apart, if a 10 percent coefficient of variation criterion is adopted. There were insufficient data for southern and western sections of the NTS to perform the geostatistical analysis. A very significant finding was that a single network design cannot be used for the entire NTS, because different areas have different variograms. Before any design can be finalized, the NTS management must specify the exposure unit area and coefficient of variation that are to be used as design criteria

  5. Characteristics and application of alanine dosimeter 'Aminogray'

    International Nuclear Information System (INIS)

    Kashiwazaki, Shigeru; Matsuyama, Shigeki; Hatta, Toshimasa; Yagyu, Hideki; Kojima, Takuji; Tanaka, Ryuichi; Morita, Yohsuke.

    1988-01-01

    Recently, accompanying the progress of nuclear power generation and space development, the evaluation of reliability for the materials and parts used under irradiation has become important. For the evaluation of reliability, the accurate grasp of radiation dose is the prerequisite. In some case, the measurement of cumulative dose in a long period in an actual environment becomes necessary. In this paper, the characteristics and application of a new dosimeter element 'Aminogray' which is suitable to the above requirement are reported. Aminogray is rodshape element made by forming alanine, a kind of amino acid, using a binder polymer, and the alanine content is 70 wt.%, and the polymer is polystyrene. An element of 3 mm diameter and 30 mm length is enclosed in a polystyrene cylinder of 4 mm thickness. This thickness was determined by considering the electronic equilibrium condition in Co-60 gamma-ray irradiation. The principle of the measurement is to determine a dose by measuring the amount of free radicals produced in alanine by radiation using ESR method. The free radicals are extremely stable, and exist for a long period, and the amount of radical production is proportional to absorbed dose. The development, characteristics and application of Aminogray are reported. (K.I.)

  6. Neutron Science 21

    International Nuclear Information System (INIS)

    Park, Sung Il; Choi, Yong Nam; Ahn, Geun Young; Lee, Hee Joo; Hong, Ji Sun; Kim, Hyo Sun

    2009-01-01

    The project aims to make the HANARO neutron beam facility a neutron research hub in the Asia-Pacific. This is a part of the effort to make the facility as productive as other neutron beam facilities in America and Europe which already operate as a hub that attracts thousands of users and produces hundreds of publications annually. The projects promotes 1) collaborative research between HANARO personnel and users, 2) internationally collaboration on various fronts of neutron science related activities. To achieve the above in an effective manner, its direction is adjusted by a series of meetings and the activities of the project are advertised to the public media. 3 domestic and 1 international collaborative research was carried out successfully in 2008 by using the HRPD. The 1st Asia-Oceania Neutron Summer School was successfully hosted in Korea. The annual report of the HANARO neutron beam facility was published. 3 advisory meetings and 6 internal meetings to promote the hub were held. Users were surveyed on their needs. The media was contacted twice in an effort to advertise the project activities. All of the above achievements are directly applicable to determining the national policy on neutron science. It is advised that the KAERI management take notice of the results to manage the facility effectively

  7. Neutron dosimetry using thermoluminescent material and KBr

    International Nuclear Information System (INIS)

    Sahyun, A.

    1979-01-01

    It is investigated the misture of natural CaF 2 and KBr as thermal neutron dosimeter compared to pure fluorite. Firstly, it was found that a 50 - 50 mixture in weight produces the highest TL sensitivity when irradiated with thermal neutrons. Second, it was shown experimentally that for TL reading within 30 to 40 hours after irradiation, the radioisotopes 42 K, 80 Br and 80 sub(m)Br contribute predominantly, while for periods longer than 4 to 5 days, 82 Br also contributes significantly. It is in question, the minimum detectable fluence is of the order of 10 9 n/cm 2 . (Autor) [pt

  8. Civilian Personnel: Career Management

    National Research Council Canada - National Science Library

    2001-01-01

    This revision; (1) Contains changes required by the establishment of a consolidated and realigned management structure for civilian personnel, manpower, and related functions in the Office of the Assistant Secretary of the Army...

  9. Tenth ORNL Personnel Dosimetry Intercomparison Study

    International Nuclear Information System (INIS)

    Swaja, R.E.; Chou, T.L.; Sims, C.S.; Greene, R.T.

    1985-03-01

    The Tenth Personnel Dosimetry Intercomparison Study was conducted at the Oak Ridge National Laboratory during April 9-11, 1984. Dosemeter badges from 31 participating organizations were mounted on 40cm Lucite phantoms and exposed to a range of dose equivalents which could be encountered during routine personnel monitoring in mixed radiation fields. The Health Physics Research Reactor served as the only source of radiation for eight of the ten irradiations which included a low (approx. 0.50 mSv) and high (approx. 10.00 mSv) neutron dose equivalent run for each of four shield conditions. Two irradiations were also conducted for which concrete- and Lucite-shield reactor irradiations were gamma-enhanced using a 137 Cs source. Results indicated that some participants had difficulty obtaining measurable indication of neutron and gamma exposures at dose equivalents less than about 0.50 mSv and 0.20 mSv, respectively. Albedo dosemeters provided the best overall accuracy and precision for the neutron measurements. Direct interaction TLD systems showed significant variation in accuracy with incident spectrum, and threshold neutron dosemeters (film and recoil track) underestimated reference values by more than 50%. Gamma dose equivalents estimated in the mixed fields were higher than reference values with TL gamma dosemeters generally yielding more accurate results than film. Under the conditions of this study in which participants had information concerning exposure conditions and radiation field characteristics prior to dosemeter evaluation, only slightly more than half of all reported results met regulatory standards for neutron and gamma accuracy. 19 refs., 2 figs., 29 tabs

  10. Neutron dosimetric measurements in shuttle and MIR

    International Nuclear Information System (INIS)

    Reitz, G.

    2001-01-01

    Detector packages consisting of thermoluminescence detectors (TLD), nuclear emulsions and plastic track detectors were exposed at identical positions inside MIR space station and on shuttle flights inside Spacelab and Spacehab during different phases of the solar cycle. The objectives of the investigations are to provide data on charge and energy spectra of heavy ions, and the contribution of events with low-energy deposit (protons, electrons, gamma, etc.) to the dose, as well as the contribution of secondaries, such as nuclear disintegration stars and neutrons. For neutron dosimetry 6 LiF (TLD600) and 7 LiF (TLD700) chips were used both of which have almost the same response to gamma rays but different response to neutrons. Neutrons in space are produced mainly in evaporation and knock-on processes with energies mainly of 1-10 MeV and up to several 100 MeV, respectively. The energy spectrum undergoes continuous changes toward greater depth in the attenuating material until an equilibrium is reached. In equilibrium, the spectrum is a wide continuum extending down to thermal energies to which the 6 LiF is sensitive. Based on the difference of absorbed doses in the 6 LiF and 7 LiF chips, thermal neutron fluxes from 1 to 2.3 cm -2 s -1 are calculated using the assumption that the maximum induced dose in TLD600 for 1 neutron cm -2 is 1.6x10 -10 Gy (Horrowitz and Freeman, Nucl. Instr. and Meth. 157 (1978) 393). It is assumed that the flux of high-energy neutrons is at least of that quantity. Tissue doses were calculated taking as a mean ambient absorbed dose per neutron 6x10 -12 Gy cm 2 (for a 10 MeV neutron). The neutron equivalent doses for the above-mentioned fluxes are 52 μGy d -1 and 120 μGy d -1 . In recent experiments, a personal neutron dosimeter was integrated into the dosimeter packages. First results of this dosimeter which is based on nuclear track detectors with converter foils are reported. For future measurements, a scintillator counter with

  11. Monte Carlo simulation experiments on box-type radon dosimeter

    International Nuclear Information System (INIS)

    Jamil, Khalid; Kamran, Muhammad; Illahi, Ahsan; Manzoor, Shahid

    2014-01-01

    Epidemiological studies show that inhalation of radon gas ( 222 Rn) may be carcinogenic especially to mine workers, people living in closed indoor energy conserved environments and underground dwellers. It is, therefore, of paramount importance to measure the 222 Rn concentrations (Bq/m 3 ) in indoors environments. For this purpose, box-type passive radon dosimeters employing ion track detector like CR-39 are widely used. Fraction of the number of radon alphas emitted in the volume of the box type dosimeter resulting in latent track formation on CR-39 is the latent track registration efficiency. Latent track registration efficiency is ultimately required to evaluate the radon concentration which consequently determines the effective dose and the radiological hazards. In this research, Monte Carlo simulation experiments were carried out to study the alpha latent track registration efficiency for box type radon dosimeter as a function of dosimeter’s dimensions and range of alpha particles in air. Two different self developed Monte Carlo simulation techniques were employed namely: (a) Surface ratio (SURA) method and (b) Ray hitting (RAHI) method. Monte Carlo simulation experiments revealed that there are two types of efficiencies i.e. intrinsic efficiency (η int ) and alpha hit efficiency (η hit ). The η int depends upon only on the dimensions of the dosimeter and η hit depends both upon dimensions of the dosimeter and range of the alpha particles. The total latent track registration efficiency is the product of both intrinsic and hit efficiencies. It has been concluded that if diagonal length of box type dosimeter is kept smaller than the range of alpha particle then hit efficiency is achieved as 100%. Nevertheless the intrinsic efficiency keeps playing its role. The Monte Carlo simulation experimental results have been found helpful to understand the intricate track registration mechanisms in the box type dosimeter. This paper explains that how radon

  12. Estimation of low-level neutron dose-equivalent rate by using extrapolation method for a curie level Am–Be neutron source

    International Nuclear Information System (INIS)

    Li, Gang; Xu, Jiayun; Zhang, Jie

    2015-01-01

    Neutron radiation protection is an important research area because of the strong radiation biological effect of neutron field. The radiation dose of neutron is closely related to the neutron energy, and the connected relationship is a complex function of energy. For the low-level neutron radiation field (e.g. the Am–Be source), the commonly used commercial neutron dosimeter cannot always reflect the low-level dose rate, which is restricted by its own sensitivity limit and measuring range. In this paper, the intensity distribution of neutron field caused by a curie level Am–Be neutron source was investigated by measuring the count rates obtained through a 3 He proportional counter at different locations around the source. The results indicate that the count rates outside of the source room are negligible compared with the count rates measured in the source room. In the source room, 3 He proportional counter and neutron dosimeter were used to measure the count rates and dose rates respectively at different distances to the source. The results indicate that both the count rates and dose rates decrease exponentially with the increasing distance, and the dose rates measured by a commercial dosimeter are in good agreement with the results calculated by the Geant4 simulation within the inherent errors recommended by ICRP and IEC. Further studies presented in this paper indicate that the low-level neutron dose equivalent rates in the source room increase exponentially with the increasing low-energy neutron count rates when the source is lifted from the shield with different radiation intensities. Based on this relationship as well as the count rates measured at larger distance to the source, the dose rates can be calculated approximately by the extrapolation method. This principle can be used to estimate the low level neutron dose values in the source room which cannot be measured directly by a commercial dosimeter. - Highlights: • The scope of the affected area for

  13. Implementation of a dosimetry service for the occupationally exposed personnel of the ESFM-IPN

    International Nuclear Information System (INIS)

    Hernandez A, Y.; Sanchez R, A. A.; Ceron R, P. V.; Rivera M, T.; Vega C, H. R.

    2014-10-01

    Because the ionizing radiations handling implies a risk for the health, is necessary to take a control of the radiation quantity that the occupationally exposed personnel receives during their daily work with ionizing radiations. For this purpose there are several types of dosimetry and companies that provide the service, but taking advantage that the Escuela Superior de Fisica y Matematicas del Instituto Politecnico Nacional (ESFM-IPN) possesses a thermoluminescence equipment (Tl) was decided to develop a personnel dosimetry service for gammas, with thermoluminescent glasses of type TLD-100. First one carries out a glasses characterization, for which the glasses were washed with a methyl alcohol solution, without rubbing them for not damaging them; after the drying, they were subjected to a temperature of 400 C during one hour and later on 100 grades C for two hours in order to erasing them the bottom information that they could have. 200 glasses were exposed to gamma radiation coming from a Cobalt 60 source property of ESFM-IPN in order to selecting the glasses that had a response whose precision was inside a standard deviation. Of this characterization 80 dosimeters were selected that had better response according to the obtained readings as well as of their shine curves. These selected TLD-100 glasses were irradiated to different dose of gamma radiation and with those readings it was built a dose response curve in m R against readings of electric load in pick Coulombs (pCou). The response curve was a direct line or calibration curve. As final exercise some glasses of this selected lot were subjected to irradiation dose not known by the person that carried out the readings and this way was proven that the results were appropriate with the calibration curve. Finally these dosimeters were placed in port-dosimeters and erased guides of dosimeters, of equipment use and dosimeters lecture were also elaborated. (Author)

  14. Modern methods to improve the accuracy in fast neutron dosimetry

    International Nuclear Information System (INIS)

    Baers, B.; Karnani, H.; Seren, T.

    1985-01-01

    In order to improve the quality of fast neutron dose estimates at the reactor pressure vessel (PV) some modern methods are presented. In addition to basic principles, some error reduction procedures are also presented based on the combined use of relative measurements, direct sample taking from the pressure vessel and the use of iron and niobium as dosimeters. The influence of large systematic errors could be significantly reduced by carrying out relative measurements. This report also presents the successful use of niobium as a dosimeter by destructive treatment of PV samples. (author)

  15. New detectors of neutron, gamma- and X-radiations

    CERN Document Server

    Lobanov, N S

    2002-01-01

    Paper presents new detectors to record absorbed doses of neutron, gamma- and X-ray radiations within 0-1500 Mrad range. DBF dosimeter is based on dibutyl phthalate. EDS dosimeter is based on epoxy (epoxide) resin, while SD 5-40 detector is based on a mixture of dibutyl phthalate and epoxy resin. Paper describes experimental techniques to calibrate and interprets the measurement results of absorbed doses for all detectors. All three detectors cover 0-30000 Mrad measured does range. The accuracy of measurements is +- 10% independent (practically) of irradiation dose rates within 20-2000 rad/s limits under 20-80 deg C temperature

  16. Application of Glycine-TTC dosimeter in gamma radiation processing facility

    International Nuclear Information System (INIS)

    Shinde, S.H.; Mondal, S.; Kulkarni, M.S.

    2018-01-01

    Glycine-TTC dosimeter was found to have a useful dose range of 5 to 30 kGy using spectro-photometric read-out method. Potential use of this dosimeter was demonstrated by measuring dose-rate in gamma chamber GC 900. The aim of the present study was to verify the performance of this dosimeter in actual industrial processing conditions encountered in radiation processing facility such as Gamma Radiation Processing Plant for Spices (GRPPS), BRIT, Vashi. Accordingly, glycine-TTC dosimeters were irradiated along with routine dosimeter viz. ceric-cerous of GRPPS and reference standard dosimeter viz. alanine EPR

  17. Ultracold neutrons

    International Nuclear Information System (INIS)

    Steenstrup, S.

    Briefly surveys recent developments in research work with ultracold neutrons (neutrons of very low velocity, up to 10 m/s at up to 10 -7 eV and 10 -3 K). Slow neutrons can be detected in an ionisation chamber filled with B 10 F 3 . Very slow neutrons can be used for investigations into the dipole moment of neutrons. Neutrons of large wave length have properties similar to those of light. The limit angle for total reflection is governed by the wave length and by the material. Total reflection can be used to filter ultracold neutrons out of the moderator material of a reactor. Total reflection can also be used to store ultracold neutrons but certain problems with storage have not yet been clarified. Slow neutrons can be made to lose speed in a neutron turbine, and come out as ultracold neutrons. A beam of ultracold neutrons could be used in a neutron microscope. (J.S.)

  18. Application of the adjoint function methodology for neutron fluence determination

    International Nuclear Information System (INIS)

    Haghighat, A.; Nanayakkara, B.; Livingston, J.; Mahgerefteh, M.; Luoma, J.

    1991-01-01

    In previous studies, the neutron fluence at a reactor pressure vessel has been estimated based on consolidation of transport theory calculations and experimental data obtained from in-vessel capsules and/or cavity dosimeters. Normally, a forward neutron transport calculation is performed for each fuel cycle and the neutron fluxes are integrated over the reactor operating time to estimate the neutron fluence. Such calculations are performed for a geometrical model which is composed of one-eighth (0 to 45 deg) of the reactor core and its surroundings; i.e., core barrel, thermal shield, downcomer, reactor vessel, cavity region, concrete wall, and instrumentation well. Because the model is large, transport theory calculations generally require a significant amount of computer memory and time; hence, more efficient methodologies such as the adjoint transport approach have been proposed. These studies, however, do not address the necessary sensitivity studies needed for adjoint function calculations. The adjoint methodology has been employed to estimate the activity of a cavity dosimeter and that of an in-vessel capsule. A sensitivity study has been performed on the mesh distribution used in and around the cavity dosimeter and the in-vessel capsule. Further, since a major portion of the detector response is due to the neutrons originated in the peripheral fuel assemblies, a study on the use of a smaller calculational model has been performed

  19. Neutron area monitor with TLD pairs

    International Nuclear Information System (INIS)

    Guzman G, K. A.; Borja H, C. G.; Valero L, C.; Hernandez D, V. M.; Vega C, H. R.

    2011-11-01

    The response of a passive neutron area monitor with pairs of thermoluminescent dosimeters has been calculated using the Monte Carlo code MCNP5. The response was calculated for one TLD 600 located at the center of a polyethylene cylinder, as moderator. When neutrons collide with the moderator lose their energy reaching the TLD with thermal energies where the ambient dose equivalent is calculated. The response was calculated for 47 monoenergetic neutron sources ranging from 1E(-9) to 20 MeV. Response was calculated using two irradiation geometries, one with an upper source and another with a lateral source. For both irradiation schemes the response was calculated with the TLDs in two positions, one parallel to the source and another perpendicular to the source. The advantage of this passive neutron monitor area is that can be used in locations with intense, pulsed and mixed radiation fields. (Author)

  20. Development of neutron calibration field using accelerators

    Energy Technology Data Exchange (ETDEWEB)

    Baba, Mamoru [Tohoku Univ., Cyclotron and Radioisotope Center, Sendai, Miyagi (Japan)

    2003-03-01

    A brief summary is given on the fast neutron calibration fields for 1) 8 keV to 15 MeV range, and 2) 30-80 MeV range. The field for 8 keV to 15 MeV range was developed at the Fast Neutron Laboratory (FNL) at Tohoku University using a 4.5 MV pulsed Dynamitron accelerator and neutron production reactions, {sup 45}Sc(p, n), {sup 7}Li(p, n), {sup 3}H(p, n), D(d, n) and T(d, n). The latter 30-80 MeV fields are setup at TIARA of Takasaki Establishment of Japan Atomic Energy Research Institute, and at Cyclotron Radio Isotope Center (CYRIC) of Tohoku University using a 90 MeV AVF cyclotron and the {sup 7}Li(p, n) reaction. These fields have been applied for various calibration of neutron spectrometers and dosimeters, and for irradiation purposes. (author)

  1. Solid-state personal dosimeter using dose conversion algorithm

    International Nuclear Information System (INIS)

    Lee, B.J.; Lee, Wanno; Cho, Gyuseong; Chang, S.Y.; Rho, S.R.

    2003-01-01

    Solid-state personal dosimeters using semiconductor detectors have been widely used because of their simplicity and real time operation. In this paper, a personal dosimeter based on a silicon PIN photodiode has been optimally designed by the Monte Carlo method and also developed. For performance test, the developed dosimeter was irradiated within the energy range between 50 keV and 1.25 MeV, the exposure dose rate between 3 mR/h and 25 R/h. The thickness of 0.2 mm Cu and 1.0 mm Al was selected as an optimal filter by simulation results. For minimizing the non-linear sensitivity on energy, dose conversion algorithm was presented, which was able to consider pulse number as well as pulse amplitude related to absorbed energies. The sensitivities of dosimeters developed by the proposed algorithm and the conventional method were compared and analyzed in detail. When dose conversion algorithm was used, the linearity of sensitivity was better about 38%. This dosimeter will be used for above 65 keV within the relative response of ±10% to 137 Cs

  2. A critical assessment of two types of personal UV dosimeters.

    Science.gov (United States)

    Seckmeyer, Gunther; Klingebiel, Marcus; Riechelmann, Stefan; Lohse, Insa; McKenzie, Richard L; Liley, J Ben; Allen, Martin W; Siani, Anna-Maria; Casale, Giuseppe R

    2012-01-01

    Doses of erythemally weighted irradiances derived from polysulphone (PS) and electronic ultraviolet (EUV) dosimeters have been compared with measurements obtained using a reference spectroradiometer. PS dosimeters showed mean absolute deviations of 26% with a maximum deviation of 44%, the calibrated EUV dosimeters showed mean absolute deviations of 15% (maximum 33%) around noon during several test days in the northern hemisphere autumn. In the case of EUV dosimeters, measurements with various cut-off filters showed that part of the deviation from the CIE erythema action spectrum was due to a small, but significant sensitivity to visible radiation that varies between devices and which may be avoided by careful preselection. Usually the method of calibrating UV sensors by direct comparison to a reference instrument leads to reliable results. However, in some circumstances the quality of measurements made with simple sensors may be over-estimated. In the extreme case, a simple pyranometer can be used as a UV instrument, providing acceptable results for cloudless skies, but very poor results under cloudy conditions. It is concluded that while UV dosimeters are useful for their design purpose, namely to estimate personal UV exposures, they should not be regarded as an inexpensive replacement for meteorological grade instruments. © 2011 Wiley Periodicals, Inc. Photochemistry and Photobiology © 2011 The American Society of Photobiology.

  3. Temperature, humidity and time., Combined effects on radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 o C for irradiation by 60 Co photons and 10-MeV electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is + 0.25 ± 0.1% per o C for the FWT-60-00 dosimeters and +0.5 ± 0.1% per o C for Riso B3 dosimeters at temperatures between 20 and 50 o C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger dependences. Whenever possible one should use dosimeters sealed in pouches under controlled intermediate humidity conditions (30-50%) or, if that is impractical, one should maintain conditions of calibration as close as possible to the conditions of use. Without that precaution, severe dosimetry errors may result. (author)

  4. An assessment of radiotherapy dosimeters based on CVD grown diamond

    International Nuclear Information System (INIS)

    Ramkumar, S.; Buttar, C.M.; Conway, J.; Whitehead, A.J.; Sussman, R.S.; Hill, G.; Walker, S.

    2001-01-01

    Diamond is potentially a very suitable material for use as a dosimeter for radiotherapy. Its radiation hardness, the near tissue equivalence and chemical inertness are some of the characteristics of diamond, which make it well suited for its application as a dosimeter. Recent advances in the synthesis of diamond by chemical vapour deposition (CVD) technology have resulted in the improvement in the quality of material and increased its suitability for radiotherapy applications. We report in this paper, the response of prototype dosimeters based on two different types (CVD1 and CVD2) of CVD diamond to X-rays. The diamond devices were assessed for sensitivity, dependence of response on dose and dose rate, and compared with a Scanditronix silicon photon diode and a PTW natural diamond dosimeter. The diamond devices of CVD1 type showed an initial increase in response with dose, which saturates after ∼6 Gy. The diamond devices of CVD2 type had a response at low fields ( 1162.8 V/cm), the CVD2-type devices showed polarisation and dose-rate dependence. The sensitivity of the CVD diamond devices varied between 82 and 1300 nC/Gy depending upon the sample type and the applied voltage. The sensitivity of CVD diamond devices was significantly higher than that of natural diamond and silicon dosimeters. The results suggest that CVD diamond devices can be fabricated for successful use in radiotherapy applications

  5. Calibration of film dosimeters by means of absorbed dose calorimeters

    International Nuclear Information System (INIS)

    Nikolaev, S.M.; Vanyushkin, B.M.; Kon'kov, N.G.

    1980-01-01

    Methods of graduating film dosimeters by means of calorimeters of absorbed doses, are considered. Graduating of film dosimeters at the energies of accelerated electrons from 4 to 10 MeV can be carried out by means of quasiadiabatic calorimeter of local absorption, the absorber thickness of which should not exceed 5-10% of Rsub(e) value, where Rsub(e) - free electron path of the given energy. In this case film is located inside the calorimeter. For graduating films with thickness not less than (0.1-0.2)Rsub(e) it is suggested to use calorimeter of full absorption; then the graduated dosimeters are located in front of the calorimeter. Graduation of films at small energies of electrons is exercised by means of a package of films, approximately Rsub(e) thick. A design of quasiadiabatic calorimeter, intended for graduating dosimeters within the energy range of electron beam from 4 to 10 MeV, is considered. The quasiadiabatic calorimeter is a thin graphite tablet with heater and thermocouple, surrounded by foam plastic thermostating case. Electricity quantity, accumulated during the radiation field pass, is measured in the case of using the quasiadiabatic calorimeter for film graduating. The results of graduating film dosimeters, obtained using film package with Rsub(e) thickness, are presented. The obtained results coincide within 5% limits with the data known beforehand [ru

  6. Lithium Blanket Module dosimetry measurements at the LOTUS 14-MeV neutron source facility

    International Nuclear Information System (INIS)

    Tsang, F.Y.; Leo, W.R.; Sahraoui, C.; Wuthrich, S.; Harker, Y.D.

    1986-01-01

    This paper describes the measurements and results of the dosimeter material reaction rates inside the Lithium Blanket Module (LBM) after irradiation by the LOTUS 14-MeV neutron source at the Ecole Polytechnique Federale de Lausanne. The measurement program has been designed to utilize sets of passive dosimeter materials in the form of foils and wires. The dosimetry materials reaction thresholds and interaction response ranges chosen for this series of measurements encompass the entire neutron spectra along the full length of the LBM fuel rods

  7. Investigation of dose distribution in mixed neutron-gamma field of boron neutron capture therapy using N isopropylacrylamide gel

    Energy Technology Data Exchange (ETDEWEB)

    Bavarmegin, Elham; Sadremomtaz, Alireza [Nuclear Science and Technology Research Institute (NSTRI), Tehran (Iran, Islamic Republic of); Khalafi, Hossein; Kasesaz, Yaser [Dept. of Physics, University of Guilan, Rasht (Iran, Islamic Republic of); Khajeali, Azim [Medical Education Research Center, Tabriz (Iran, Islamic Republic of)

    2017-02-15

    Gel dosimeters have unique advantages in comparison with other dosimeters. Until now, these gels have been used in different radiotherapy techniques as a reliable dosimetric tool. Because dose distribution measurement is an important factor for appropriate treatment planning in different radiotherapy techniques, in this study, we evaluated the ability of the N-isopropylacrylamide (NIPAM) polymer gel to record the dose distribution resulting from the mixed neutron-gamma field of boron neutron capture therapy (BNCT). In this regard, a head phantom containing NIPAM gel was irradiated using the Tehran Research Reactor BNCT beam line, and then by a magnetic resonance scanner. Eventually, the R2 maps were obtained in different slices of the phantom by analyzing T2-weighted images. The results show that NIPAM gel has a suitable potential for recording three-dimensional dose distribution in mixed neutron-gamma field dosimetry.

  8. LANSCE personnel access control system (PACS)

    International Nuclear Information System (INIS)

    Sturrock, J.C.; Gallegos, F.R.; Hall, M.J.

    1997-01-01

    The Radiation Security System (RSS) at the Los Alamos Neutron Science Center (LANSCE) provides personnel protection from prompt radiation due to accelerated beam. The Personnel Access Control System (PACS) is a component of the RSS that is designed to prevent personnel access to areas where prompt radiation is a hazard. PACS was designed to replace several older personnel safety systems (PSS) with a single modem unified design. Lessons learned from the operation over the last 20 years were incorporated into a redundant sensor, single-point failure safe, fault tolerant, and tamper-resistant system that prevents access to the beam areas by controlling the access keys and beam stoppers. PACS uses a layered philosophy to the physical and electronic design. The most critical assemblies are battery backed up, relay logic circuits; less critical devices use Programmable Logic Controllers (PLCs) for timing functions and communications. Outside reviewers have reviewed the operational safety of the design. The design philosophy, lessons learned, hardware design, software design, operation, and limitations of the device are described

  9. The current status of eye lens dose measurement in interventional cardiology personnel in Thailand.

    Science.gov (United States)

    Krisanachinda, Anchali; Srimahachota, Suphot; Matsubara, Kosuke

    2017-06-01

    Workers involved in interventional cardiology procedures receive high eye lens doses if radiation protection tools are not properly utilized. Currently, there is no suitable method for routine measurement of eye dose. In Thailand, the eye lens equivalent doses in terms of Hp(3) of the interventional cardiologists, nurses, and radiographers participating in interventional cardiology procedures have been measured at 12 centers since 2015 in the pilot study. The optically stimulated luminescence (OSL) dosimeter was used for measurement of the occupational exposure and the eye lens dose of 42 interventional cardiology personnel at King Chulalongkorn Memorial Hospital as one of the pilot centers. For all personnel, it is recommended that a first In Light OSL badge is placed at waist level and under the lead apron for determination of Hp(10); a second badge is placed at the collar for determination of Hp(0.07) and estimation of Hp(3). Nano Dots OSL dosimeter has been used as an eye lens dosimeter for 16 interventional cardiology personnel, both with and without lead-glass eyewear. The mean effective dose at the body, equivalent dose at the collar, and estimated eye lens dose were 0.801, 5.88, and 5.70 mSv per year, respectively. The mean eye lens dose measured by the Nano Dots dosimeter was 8.059 mSv per year on the left eye and 3.552 mSv per year on the right eye. Two of 16 interventional cardiologists received annual eye lens doses on the left side without lead glass that were higher than 20 mSv per year, the new eye lens dose limit as recommended by ICRP with the risk of eye lens opacity and cataract.

  10. Considerations in the design of an improved transportable neutron spectrometer

    CERN Document Server

    Williams, A M; Brushwood, J M; Beeley, P A

    2002-01-01

    The Transportable Neutron Spectrometer (TNS) has been used by the Ministry of Defence for over 15 years to characterise neutron fields in workplace environments and provide local correction factors for both area and personal dosimeters. In light of advances in neutron spectrometry, a programme to evaluate and improve TNS has been initiated. This paper describes TNS, presents its operation in known radioisotope fields and in a reactor environment. Deficiencies in the operation of the instrument are highlighted, together with proposals for updating the response functions and spectrum unfolding methodologies.

  11. Photon energy response of an aluminum oxide TLD environmental dosimeter

    International Nuclear Information System (INIS)

    Olsher, R.H.

    1992-01-01

    Because of aluminum oxide's significant advantage in sensitivity (about a factor of 30) over LiF, minimal fading characteristics and ease of processing, aluminum oxide thermoluminescent dosimeters (TLDS) are being phased in at Los alamos for environmental monitoring of photon radiation. The new environmental dosimeter design consists of a polyethylene holder, about 0. 5 cm thick, loaded with a stack of four aluminum oxide TLD chips, each 1 mm thick and 5 mm in diameter. As part of the initial evaluation of the new design, the photon energy response of the dosimeter was calculated over the range from 10 keV to 1 MeV. Specific goals of the analysis included the determination of individual chip response in the stack, assessment of the response variation due to TLD material (i.e., LiF versus A1 2 O 3 ), and the effect of copper filtration in flattening the response

  12. Training of maintenance personnel

    International Nuclear Information System (INIS)

    Rabouhams, J.

    1986-01-01

    This lecture precises the method and means developed by EDF to ensure the training of maintenance personnel according to their initial educational background and their experience. The following points are treated: General organization of the training for maintenance personnel in PWR and GCR nuclear power stations and in Creys Malville fast breeder reactor; Basic nuclear training and pedagogical aids developed for this purpose; Specific training and training provided by contractors; complementary training taking into account the operation experience and feedback; Improvement of velocity, competence and safety during shut-down operations by adapted training. (orig.)

  13. Passive dosimetry: introduction of a new dosimeter based on OSL technology

    International Nuclear Information System (INIS)

    Archambault, V.; Le Roy, G.; Prugnaud, B.

    2005-01-01

    A new passive dosimeter based on OSL technology has been introduced on the French market. In this article are described: the technology and the material on which this new detector relied, the dosimeter itself. (author)

  14. Angular dependence of the nanoDot OSL dosimeter

    International Nuclear Information System (INIS)

    Kerns, James R.; Kry, Stephen F.; Sahoo, Narayan; Followill, David S.; Ibbott, Geoffrey S.

    2011-01-01

    Purpose: Optically stimulated luminescent detectors (OSLDs) are quickly gaining popularity as passive dosimeters, with applications in medicine for linac output calibration verification, brachytherapy source verification, treatment plan quality assurance, and clinical dose measurements. With such wide applications, these dosimeters must be characterized for numerous factors affecting their response. The most abundant commercial OSLD is the InLight/OSL system from Landauer, Inc. The purpose of this study was to examine the angular dependence of the nanoDot dosimeter, which is part of the InLight system. Methods: Relative dosimeter response data were taken at several angles in 6 and 18 MV photon beams, as well as a clinical proton beam. These measurements were done within a phantom at a depth beyond the build-up region. To verify the observed angular dependence, additional measurements were conducted as well as Monte Carlo simulations in MCNPX. Results: When irradiated with the incident photon beams parallel to the plane of the dosimeter, the nanoDot response was 4% lower at 6 MV and 3% lower at 18 MV than the response when irradiated with the incident beam normal to the plane of the dosimeter. Monte Carlo simulations at 6 MV showed similar results to the experimental values. Examination of the results in Monte Carlo suggests the cause as partial volume irradiation. In a clinical proton beam, no angular dependence was found. Conclusions: A nontrivial angular response of this OSLD was observed in photon beams. This factor may need to be accounted for when evaluating doses from photon beams incident from a variety of directions.

  15. Exposure potential to neutrons of the Brazilian workers

    International Nuclear Information System (INIS)

    Martins, Marcelo M.; Almeida, Carlos E. de

    1995-01-01

    Nowadays, there are 222 neutron radioactive sources registered in use in Brazil, in addition to several neutron fields, power and research reactors and neutron generators. Secondary neutron fields can also be generated in particle accelerators by nuclear reactions with its shielding, the experimental set up and the conversion target. These neutron fields are very different, not only in their spectra but also in their fluences. Around 200 radiation workers are monthly monitored since 1983 by an albedo system. Of the evaluated dosimeters 4% only have shown neutron doses, being 13 mSv the maximum measured value. Most of these doses were received by workers of the oil-well logging and research activities users of 241 Am-Be sources. (author). 7 refs., 3 tabs

  16. Comparison between two kind of power circuits for personal dosimeter

    International Nuclear Information System (INIS)

    Liu Zhengshan; Deng Changming; Guo Zhanjie

    2002-01-01

    Personal Dosimeter is commonly requested using battery for its power supply, and hope the battery life is long. Also with the fall of battery voltage, some performance of instrument as well as drop, Reasonable supply design can protract the battery life. The author introduces two method: power supply with battery directly and supply used power chip conversion. Combine personal dosimeter, the authors carried comparison for battery life, power consumption, cost and volume. Based on the comparison result and instrument fact request, one can choose method of power circuit

  17. Locations of criticality alarms and nuclear accident dosimeters at Hanford

    International Nuclear Information System (INIS)

    1992-08-01

    Hanford facilities that contain fissionable materials capable of achieving critical mass are monitored with nuclear accident dosimeters (NADS) in compliance with the requirements of DOE Order 5480.11, Chapter XI, Section 4.c. (DOE 1988). The US Department of Energy (DOE) Richland Field Office (RL) has assigned the responsibility for maintaining and evaluating the Hanford NAD system to the Instrumentation and External Dosimetry (I ampersand ED) Section of Pacific Northwest Laboratory's (PNL's) Health Physics Department. This manual provides a description of the Hanford NAD, criteria and instructions for proper NAD placement, and the locations of these dosimeters onsite

  18. Current personnel dosimetry practices at DOE facilities

    International Nuclear Information System (INIS)

    Fix, J.J.

    1981-05-01

    Only three parameters were included in the personnel occupational exposure records by all facilities. These are employee name, social security number, and whole body dose. Approximate percentages of some other parameters included in the record systems are sex (50%), birthdate (90%), occupation (26%), previous employer radiation exposure (74%), etc. Statistical analysis of the data for such parameters as sex versus dose distribution, age versus dose distribution, cumulative lifetime dose, etc. was apparently seldom done. Less than 50% of the facilities reported having formal documentation for either the dosimeter, records system, or reader. Slightly greater than 50% of facilities reported having routine procedures in place. These are considered maximum percentages because some respondents considered computer codes as formal documentation. The repository receives data from DOE facilities regarding the (a) distribution of annual whole body doses, (b) significant internal depositions, and (c) individual doses upon termination. It is expected that numerous differences exist in the dose data submitted by the different facilities. Areas of significant differences would likely include the determination of non-measurable doses, the methods used to determine previous employer radiation dose, the methods of determining cumulative radiation dose, and assessment of internal doses. Undoubtedly, the accuracy of the different dosimetry systems, especially at low doses, is very important to the credibility of data summaries (e.g., man-rem) provided by the repository

  19. Neutron/gamma dose separation by the multiple-ion-chamber technique

    International Nuclear Information System (INIS)

    Goetsch, S.J.

    1983-01-01

    Many mixed n/γ dosimetry systems rely on two dosimeters, one composed of a tissue-equivalent material and the other made from a non-hydrogenous material. The paired chamber technique works well in fields of neutron radiation nearly identical in spectral composition to that in which the dosimeters were calibrated. However, this technique is drastically compromised in phantom due to the degradation of the neutron spectrum. The three-dosimeter technique allows for the fall-off in neutron sensitivity of the two non-hydrogenous dosimeters. Precise and physically meaningful results were obtained with this technique with a D-T source in air and in phantom and with simultaneous D-T neutron and 60 Co gamma ray irradiation in air. The MORSE-CG coupled n/γ three-dimensional Monte Carlo code was employed to calculate neutron and gamma doses in a water phantom. Gamma doses calculated in phantom with this code were generally lower than corresponding ion chamber measurements. This can be explained by the departure of irradiation conditions from ideal narrow-beam geometry. 97 references

  20. Personnel photographic film dosimetry

    International Nuclear Information System (INIS)

    Keirim-Markus, I.B.

    1981-01-01

    Technology of personnel photographic film dosimetry (PPD) based on the photographic effect of ionizing radiation is described briefly. Kinds of roentgen films used in PPD method are enumerated, compositions of a developer and fixing agents for these films are given [ru