WorldWideScience

Sample records for maximum count rates

  1. TU-FG-209-03: Exploring the Maximum Count Rate Capabilities of Photon Counting Arrays Based On Polycrystalline Silicon

    Energy Technology Data Exchange (ETDEWEB)

    Liang, A K; Koniczek, M; Antonuk, L E; El-Mohri, Y; Zhao, Q [University of Michigan, Ann Arbor, MI (United States)

    2016-06-15

    Purpose: Photon counting arrays (PCAs) offer several advantages over conventional, fluence-integrating x-ray imagers, such as improved contrast by means of energy windowing. For that reason, we are exploring the feasibility and performance of PCA pixel circuitry based on polycrystalline silicon. This material, unlike the crystalline silicon commonly used in photon counting detectors, lends itself toward the economic manufacture of radiation tolerant, monolithic large area (e.g., ∼43×43 cm2) devices. In this presentation, exploration of maximum count rate, a critical performance parameter for such devices, is reported. Methods: Count rate performance for a variety of pixel circuit designs was explored through detailed circuit simulations over a wide range of parameters (including pixel pitch and operating conditions) with the additional goal of preserving good energy resolution. The count rate simulations assume input events corresponding to a 72 kVp x-ray spectrum with 20 mm Al filtration interacting with a CZT detector at various input flux rates. Output count rates are determined at various photon energy threshold levels, and the percentage of counts lost (e.g., due to deadtime or pile-up) is calculated from the ratio of output to input counts. The energy resolution simulations involve thermal and flicker noise originating from each circuit element in a design. Results: Circuit designs compatible with pixel pitches ranging from 250 to 1000 µm that allow count rates over a megacount per second per pixel appear feasible. Such rates are expected to be suitable for radiographic and fluoroscopic imaging. Results for the analog front-end circuitry of the pixels show that acceptable energy resolution can also be achieved. Conclusion: PCAs created using polycrystalline silicon have the potential to offer monolithic large-area detectors with count rate performance comparable to those of crystalline silicon detectors. Further improvement through detailed circuit

  2. Compton suppression gamma-counting: The effect of count rate

    Science.gov (United States)

    Millard, H.T.

    1984-01-01

    Past research has shown that anti-coincidence shielded Ge(Li) spectrometers enhanced the signal-to-background ratios for gamma-photopeaks, which are situated on high Compton backgrounds. Ordinarily, an anti- or non-coincidence spectrum (A) and a coincidence spectrum (C) are collected simultaneously with these systems. To be useful in neutron activation analysis (NAA), the fractions of the photopeak counts routed to the two spectra must be constant from sample to sample to variations must be corrected quantitatively. Most Compton suppression counting has been done at low count rate, but in NAA applications, count rates may be much higher. To operate over the wider dynamic range, the effect of count rate on the ratio of the photopeak counts in the two spectra (A/C) was studied. It was found that as the count rate increases, A/C decreases for gammas not coincident with other gammas from the same decay. For gammas coincident with other gammas, A/C increases to a maximum and then decreases. These results suggest that calibration curves are required to correct photopeak areas so quantitative data can be obtained at higher count rates. ?? 1984.

  3. Maximum entropy deconvolution of low count nuclear medicine images

    International Nuclear Information System (INIS)

    McGrath, D.M.

    1998-12-01

    Maximum entropy is applied to the problem of deconvolving nuclear medicine images, with special consideration for very low count data. The physics of the formation of scintigraphic images is described, illustrating the phenomena which degrade planar estimates of the tracer distribution. Various techniques which are used to restore these images are reviewed, outlining the relative merits of each. The development and theoretical justification of maximum entropy as an image processing technique is discussed. Maximum entropy is then applied to the problem of planar deconvolution, highlighting the question of the choice of error parameters for low count data. A novel iterative version of the algorithm is suggested which allows the errors to be estimated from the predicted Poisson mean values. This method is shown to produce the exact results predicted by combining Poisson statistics and a Bayesian interpretation of the maximum entropy approach. A facility for total count preservation has also been incorporated, leading to improved quantification. In order to evaluate this iterative maximum entropy technique, two comparable methods, Wiener filtering and a novel Bayesian maximum likelihood expectation maximisation technique, were implemented. The comparison of results obtained indicated that this maximum entropy approach may produce equivalent or better measures of image quality than the compared methods, depending upon the accuracy of the system model used. The novel Bayesian maximum likelihood expectation maximisation technique was shown to be preferable over many existing maximum a posteriori methods due to its simplicity of implementation. A single parameter is required to define the Bayesian prior, which suppresses noise in the solution and may reduce the processing time substantially. Finally, maximum entropy deconvolution was applied as a pre-processing step in single photon emission computed tomography reconstruction of low count data. Higher contrast results were

  4. Relationship of milking rate to somatic cell count.

    Science.gov (United States)

    Brown, C A; Rischette, S J; Schultz, L H

    1986-03-01

    Information on milking rate, monthly bucket somatic cell counts, mastitis treatment, and milk production was obtained from 284 lactations of Holstein cows separated into three lactation groups. Significant correlations between somatic cell count (linear score) and other parameters included production in lactation 1 (-.185), production in lactation 2 (-.267), and percent 2-min milk in lactation 2 (.251). Somatic cell count tended to increase with maximum milking rate in all lactations, but correlations were not statistically significant. Twenty-nine percent of cows with milking rate measurements were treated for clinical mastitis. Treated cows in each lactation group produced less milk than untreated cows. In the second and third lactation groups, treated cows had a shorter total milking time and a higher percent 2-min milk than untreated cows, but differences were not statistically significant. Overall, the data support the concept that faster milking cows tend to have higher cell counts and more mastitis treatments, particularly beyond first lactation. However, the magnitude of the relationship was small.

  5. Quantitative Compton suppression spectrometry at elevated counting rates

    International Nuclear Information System (INIS)

    Westphal, G.P.; Joestl, K.; Schroeder, P.; Lauster, R.; Hausch, E.

    1999-01-01

    For quantitative Compton suppression spectrometry the decrease of coincidence efficiency with counting rate should be made negligible to avoid a virtual increase of relative peak areas of coincident isomeric transitions with counting rate. To that aim, a separate amplifier and discriminator has been used for each of the eight segments of the active shield of a new well-type Compton suppression spectrometer, together with an optimized, minimum dead-time design of the anticoincidence logic circuitry. Chance coincidence losses in the Compton suppression spectrometer are corrected instrumentally by comparing the chance coincidence rate to the counting rate of the germanium detector in a pulse-counting Busy circuit (G.P. Westphal, J. Rad. Chem. 179 (1994) 55) which is combined with the spectrometer's LFC counting loss correction system. The normally not observable chance coincidence rate is reconstructed from the rates of germanium detector and scintillation detector in an auxiliary coincidence unit, after the destruction of true coincidence by delaying one of the coincidence partners. Quantitative system response has been tested in two-source measurements with a fixed reference source of 60 Co of 14 kc/s, and various samples of 137 Cs, up to aggregate counting rates of 180 kc/s for the well-type detector, and more than 1400 kc/s for the BGO shield. In these measurements, the net peak areas of the 1173.3 keV line of 60 Co remained constant at typical values of 37 000 with and 95 000 without Compton suppression, with maximum deviations from the average of less than 1.5%

  6. Delta count-rate monitoring system

    International Nuclear Information System (INIS)

    Van Etten, D.; Olsen, W.A.

    1985-01-01

    A need for a more effective way to rapidly search for gamma-ray contamination over large areas led to the design and construction of a very sensitive gamma detection system. The delta count-rate monitoring system was installed in a four-wheel-drive van instrumented for environmental surveillance and accident response. The system consists of four main sections: (1) two scintillation detectors, (2) high-voltage power supply amplifier and single-channel analyzer, (3) delta count-rate monitor, and (4) count-rate meter and recorder. The van's 6.5-kW generator powers the standard nuclear instrument modular design system. The two detectors are mounted in the rear corners of the van and can be run singly or jointly. A solid-state bar-graph count-rate meter mounted on the dashboard can be read easily by both the driver and passenger. A solid-state strip chart recorder shows trends and provides a permanent record of the data. An audible alarm is sounded at the delta monitor and at the dashboard count-rate meter if a detected radiation level exceeds the set background level by a predetermined amount

  7. Bayesian analysis of energy and count rate data for detection of low count rate radioactive sources.

    Science.gov (United States)

    Klumpp, John; Brandl, Alexander

    2015-03-01

    A particle counting and detection system is proposed that searches for elevated count rates in multiple energy regions simultaneously. The system analyzes time-interval data (e.g., time between counts), as this was shown to be a more sensitive technique for detecting low count rate sources compared to analyzing counts per unit interval (Luo et al. 2013). Two distinct versions of the detection system are developed. The first is intended for situations in which the sample is fixed and can be measured for an unlimited amount of time. The second version is intended to detect sources that are physically moving relative to the detector, such as a truck moving past a fixed roadside detector or a waste storage facility under an airplane. In both cases, the detection system is expected to be active indefinitely; i.e., it is an online detection system. Both versions of the multi-energy detection systems are compared to their respective gross count rate detection systems in terms of Type I and Type II error rates and sensitivity.

  8. Count rate effect in proportional counters

    International Nuclear Information System (INIS)

    Bednarek, B.

    1980-01-01

    A new concept is presented explaining changes in spectrometric parameters of proportional counters which occur due to varying count rate. The basic feature of this concept is that the gas gain of the counter remains constant in a wide range of count rate and that the decrease in the pulse amplitude and the detorioration of the energy resolution observed are the results of changes in the shape of original current pulses generated in the active volume of the counter. In order to confirm the validity of this statement, measurements of the gas amplification factor have been made in a wide count rate range. It is shown that above a certain critical value the gas gain depends on both the operating voltage and the count rate. (author)

  9. Fast radio burst event rate counts - I. Interpreting the observations

    Science.gov (United States)

    Macquart, J.-P.; Ekers, R. D.

    2018-02-01

    The fluence distribution of the fast radio burst (FRB) population (the `source count' distribution, N (>F) ∝Fα), is a crucial diagnostic of its distance distribution, and hence the progenitor evolutionary history. We critically reanalyse current estimates of the FRB source count distribution. We demonstrate that the Lorimer burst (FRB 010724) is subject to discovery bias, and should be excluded from all statistical studies of the population. We re-examine the evidence for flat, α > -1, source count estimates based on the ratio of single-beam to multiple-beam detections with the Parkes multibeam receiver, and show that current data imply only a very weak constraint of α ≲ -1.3. A maximum-likelihood analysis applied to the portion of the Parkes FRB population detected above the observational completeness fluence of 2 Jy ms yields α = -2.6_{-1.3}^{+0.7 }. Uncertainties in the location of each FRB within the Parkes beam render estimates of the Parkes event rate uncertain in both normalizing survey area and the estimated post-beam-corrected completeness fluence; this uncertainty needs to be accounted for when comparing the event rate against event rates measured at other telescopes.

  10. PVO VENUS ONMS BROWSE SUPRTHRML ION MAX COUNT RATE 12S V1.0

    Data.gov (United States)

    National Aeronautics and Space Administration — This data set contains PVO Neutral Mass Spectrometer superthermal ion data. Each record contains the maximum count rate per second in a 12 second period beginning...

  11. Count-rate analysis from clinical scans in PET with LSO detectors

    International Nuclear Information System (INIS)

    Bonutti, F.; Cattaruzzi, E.; Cragnolini, E.; Floreani, M.; Foti, C.; Malisan, M. R.; Moretti, E.; Geatti, O.; Padovani, R.

    2008-01-01

    The purpose of optimising the acquisition parameters in positron emission tomography is to improve the quality of the diagnostic images. Optimisation can be done by maximising the noise equivalent count rate (NECR) that in turn depends on the coincidence rate. For each bed position the scanner records coincidences and singles rates. For each patient, the true, random and scattered coincidences as functions of the single count rate(s) are determined by fitting the NEMA (National Electrical Manufacturers Association) 70 cm phantom count rate curves to measured clinical points. This enables analytical calculation of the personalised PNECR [pseudo NECR(s)] curve, linked to the NECR curve. For central bed positions, missing activity of ∼70% is estimated to get maximum PNECR (PNECR max ), but the improvement in terms of signal-to-noise ratio would be ∼15%. The correlation between patient weight and PNECR max is also estimated to determine the optimal scan duration of a single bed position as a function of patient weight at the same PNEC. Normalising the counts at PNECR max for the 70 kg patient, the bed duration for a 90 kg patient should be 230 s, which is ∼30% longer. Although the analysis indicates that the fast scanner electronics allow using higher administered activities, this would involve poor improvement in terms of NECR. Instead, attending to higher bed duration for heavier patients may be more useful. (authors)

  12. A rotation-symmetric, position-sensitive annular detector for maximum counting rates

    International Nuclear Information System (INIS)

    Igel, S.

    1993-12-01

    The Germanium Wall is a semiconductor detector system containing up to four annular position sensitive ΔE-detectors from high purity germanium (HPGe) planned to complement the BIG KARL spectrometer in COSY experiments. The first diode of the system, the Quirl-detector, has a two dimensional position sensitive structure defined by 200 Archimedes' spirals on each side with opposite orientation. In this way about 40000 pixels are defined. Since each spiral element detects almost the same number of events in an experiment the whole system can be optimized for maximal counting rates. This paper describes a test setup for a first prototype of the Quirl-detector and the results of test measurements with an α-source. The detector current and the electrical separation of the spiral elements were measured. The splitting of signals due to the spread of charge carriers produced by an incident ionizing particle on several adjacent elements was investigated in detail and found to be twice as high as expected from calculations. Its influence on energy and position resolution is discussed. Electronic crosstalk via signal wires and the influence of noise from the magnetic spectrometer has been tested under experimental conditions. Additionally, vacuum feedthroughs based on printed Kapton foils pressed between Viton seals were fabricated and tested successfully concerning their vacuum and thermal properties. (orig.)

  13. A counting-card circuit based on PCI bus

    International Nuclear Information System (INIS)

    Shi Jing; Li Yong; Chinese Academy of Sciences, Lanzhou; Su Hong; Dong Chengfu; Li Xiaogang; Ma Xiaoli

    2004-01-01

    A counting-card circuit based on PCI bus that we developed recently used for advanced personal computer will be introduced in this paper briefly. The maximum count capacity of this counting-card is 10 9 -1, ranging from 0 to 999 999 999, the maximum counting time range, 1 x 10 6 s, can be set in 1 cycle, the maximum counting rate is 20 MHz for positive input. (authors)

  14. Evaluation of misplaced event count rate using a scintillation camera

    International Nuclear Information System (INIS)

    Yanagimoto, Shin-ichi; Tomomitsu, Tatsushi; Muranaka, Akira

    1985-01-01

    Misplaced event count rates were evaluated using an acryl scatter body of various thickness and a gamma camera. The count rate in the region of interest (ROI) within the camera view field, which was thought to represent part of the misplaced event count rate, increased as the thickness of the scatter body was increased to 5 cm, followed by a steep decline in the count rate. On the other hand, the ratio of the count rate in the ROI to the total count rate continuously increased as the thickness of the scatter body was increased. As the thickness of the scatter body was increased, the count rates increased, and the increments of increase were greater in the lower energy region of the photopeak than in the higher energy region. In ranges energy other than the photopeak, the influence of the scatter body on the count rate in the ROI was the greatest at 76 keV, which was the lowest energy we examined. (author)

  15. Fast maximum likelihood estimation of mutation rates using a birth-death process.

    Science.gov (United States)

    Wu, Xiaowei; Zhu, Hongxiao

    2015-02-07

    Since fluctuation analysis was first introduced by Luria and Delbrück in 1943, it has been widely used to make inference about spontaneous mutation rates in cultured cells. Under certain model assumptions, the probability distribution of the number of mutants that appear in a fluctuation experiment can be derived explicitly, which provides the basis of mutation rate estimation. It has been shown that, among various existing estimators, the maximum likelihood estimator usually demonstrates some desirable properties such as consistency and lower mean squared error. However, its application in real experimental data is often hindered by slow computation of likelihood due to the recursive form of the mutant-count distribution. We propose a fast maximum likelihood estimator of mutation rates, MLE-BD, based on a birth-death process model with non-differential growth assumption. Simulation studies demonstrate that, compared with the conventional maximum likelihood estimator derived from the Luria-Delbrück distribution, MLE-BD achieves substantial improvement on computational speed and is applicable to arbitrarily large number of mutants. In addition, it still retains good accuracy on point estimation. Published by Elsevier Ltd.

  16. Relationship between salivary flow rates and Candida albicans counts.

    Science.gov (United States)

    Navazesh, M; Wood, G J; Brightman, V J

    1995-09-01

    Seventy-one persons (48 women, 23 men; mean age, 51.76 years) were evaluated for salivary flow rates and Candida albicans counts. Each person was seen on three different occasions. Samples of unstimulated whole, chewing-stimulated whole, acid-stimulated parotid, and candy-stimulated parotid saliva were collected under standardized conditions. An oral rinse was also obtained and evaluated for Candida albicans counts. Unstimulated and chewing-stimulated whole flow rates were negatively and significantly (p Candida counts. Unstimulated whole saliva significantly (p Candida counts of 0 versus or = 500 count. Differences in stimulated parotid flow rates were not significant among different levels of Candida counts. The results of this study reveal that whole saliva is a better predictor than parotid saliva in identification of persons with high Candida albicans counts.

  17. 7 CFR 3565.210 - Maximum interest rate.

    Science.gov (United States)

    2010-01-01

    ... 7 Agriculture 15 2010-01-01 2010-01-01 false Maximum interest rate. 3565.210 Section 3565.210... AGRICULTURE GUARANTEED RURAL RENTAL HOUSING PROGRAM Loan Requirements § 3565.210 Maximum interest rate. The interest rate for a guaranteed loan must not exceed the maximum allowable rate specified by the Agency in...

  18. A scintillation detector signal processing technique with active pileup prevention for extending scintillation count rates

    International Nuclear Information System (INIS)

    Wong, W.H.; Li, H.

    1998-01-01

    A new method for processing signals from scintillation detectors is proposed for very high count-rate situations where multiple-event pileups are the norm. This method is designed to sort out and recover every impinging event from multiple-event pileups while maximizing the collection of scintillation signal for every event to achieve optimal accuracy in determining the energy of the event. For every detected event, this method cancels the remnant signals from previous events, and excludes the pileup of signals from following events. With this technique, pileup events can be recovered and the energy of every recovered event can be optimally measured despite multiple pileups. A prototype circuit demonstrated that the maximum count rates have been increased by more than 10 times, comparing to the standard pulse-shaping method, while the energy resolution is as good as that of the pulse shaping (or the fixed integration) method at normal count rates. At 2 x 10 6 events/sec for NaI(Tl), the true counts acquired are 3 times more than the delay-line clipping method (commonly used in fast processing designs) due to events recovered from pileups. Pulse-height spectra up to 3.5 x 10 6 events/sec have been studied

  19. A multiwire proportional counter for very high counting rates

    International Nuclear Information System (INIS)

    Barbosa, A.F.; Guedes, G.P.; Tamura, E.; Pepe, I.M.; Oliveira, N.B.

    1997-12-01

    Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10 6 counts/s per wire can be reached without critical loss in the true versus measured linearity relation. Results obtained with a detector containing 30 active wires (2 mm pitch) are presented. To each wire is associated a fast pre-amplifier and a discriminator channel. Global counting rates in excess to 10 7 events/s are reported. Data acquisition systems are described for 1D (real time) and 2D (off-line) position sensitive detection systems. (author)

  20. Investigating the impact of LSO on the count rate of wholebody PET tomographs

    International Nuclear Information System (INIS)

    Douglas, J.L.; Moisan, C.; Rogers, J.G.

    1996-05-01

    We investigated the impact of using detectors made of lutetium oxyorthosilicate (LSO) on the count rate performances of wholebody PET tomographs. To that end, we used a single computational model that predicts the prompt and random contributions to the total event rate in septaless PET tomographs. Dead time factors at all stages of a typical event acquisition stream are calculated from specified values of their respective processing clock cycle. We validated our approach by fitting the true, random and multiple count rates measured with the ECAT-953B and the EXACT HR PLUS scanners for a standard 20 x 20 cm cylindrical phantom. We then investigated the implications of using position encoding detectors made of LSO in the EXACT HR PLUS scanner geometry. The results in indicate that only replacing BGO by the faster LSO incurs no appreciable change in the maximum noise-equivalent-count (NEC) rate of the scanner. However, one could realistically increase the NEC by a factor 2.5 using a 4 nsec coincidence window width with the detector processing front-end operating on a 128 nsec clock cycle. Further reducing the coincidence window width to 600 psec and the front-end clock cycle to 64 nsec leads to an increase of the NEC by a factor 7.5. To sustain the operation of an HR Plus with these specifications, the saturation bandwidth of the data acquisition system would have to be increased to no more than 4.5 MHz. (authors)

  1. A multiwire proportional counter for very high counting rates

    Energy Technology Data Exchange (ETDEWEB)

    Barbosa, A F; Guedes, G P [Centro Brasileiro de Pesquisas Fisicas (CBPF), Rio de Janeiro, RJ (Brazil); Tamura, E [Laboratorio Nacional de Luz Sincrotron (LNLS), Campinas, SP (Brazil); Pepe, I M; Oliveira, N B [Bahia Univ., Salvador, BA (Brazil). Inst. de Fisica

    1997-12-01

    Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10{sup 6} counts/s per wire can be reached without critical loss in the true versus measured linearity relation. Results obtained with a detector containing 30 active wires (2 mm pitch) are presented. To each wire is associated a fast pre-amplifier and a discriminator channel. Global counting rates in excess to 10{sup 7} events/s are reported. Data acquisition systems are described for 1D (real time) and 2D (off-line) position sensitive detection systems. (author) 13 refs., 6 figs.

  2. Correction to the count-rate detection limit and sample/blank time-allocation methods

    International Nuclear Information System (INIS)

    Alvarez, Joseph L.

    2013-01-01

    A common form of count-rate detection limits contains a propagation of uncertainty error. This error originated in methods to minimize uncertainty in the subtraction of the blank counts from the gross sample counts by allocation of blank and sample counting times. Correct uncertainty propagation showed that the time allocation equations have no solution. This publication presents the correct form of count-rate detection limits. -- Highlights: •The paper demonstrated a proper method of propagating uncertainty of count rate differences. •The standard count-rate detection limits were in error. •Count-time allocation methods for minimum uncertainty were in error. •The paper presented the correct form of the count-rate detection limit. •The paper discussed the confusion between count-rate uncertainty and count uncertainty

  3. Effective dark count rate reduction by modified SPAD gating circuit

    Energy Technology Data Exchange (ETDEWEB)

    Prochazka, Ivan; Blazej, Josef, E-mail: blazej@fjfi.cvut.cz; Kodet, Jan

    2015-07-01

    For our main application of single photon counting avalanche detectors in focus – laser ranging of space objects and laser time transfer – the ultimate requirements are relatively large and homogeneous active area having a diameter of 100 to 200 µm and a sub-picosecond stability of timing. The detector dark count rate and after-pulsing probability are parameters of relatively lower, but not negligible importance. In presented paper we will focused on them. We have developed a new active quenching and gating scheme which can reduce afterpulsing effect and hence also effective dark count rate at lower temperature. In satellite laser ranging system the effective dark count rate was reduced more than 35 times. This improvement will contribute in increasing the data yield and hence to increase precision and productivity. - Highlights: • Signal and quenching path in a control circuit stayed unaffected by gating. • The detector package optimized for laser time transfer systems is considered. • After-pulsing effect is reduced by a modification of the use of gate signal. • The dark count rate is reduced for gate rates of the order of units of kHz.

  4. A high count rate position decoding and energy measuring method for nuclear cameras using Anger logic detectors

    International Nuclear Information System (INIS)

    Wong, W.H.; Li, H.; Uribe, J.

    1998-01-01

    A new method for processing signals from Anger position-sensitive detectors used in gamma cameras and PET is proposed for very high count-rate imaging where multiple-event pileups are the norm. This method is designed to sort out and recover every impinging event from multiple-event pileups while maximizing the collection of scintillation signal for every event to achieve optimal accuracy in the measurement of energy and position. For every detected event, this method cancels the remnant signals from previous events, and excludes the pileup of signals from following events. The remnant subtraction is exact even for multiple pileup events. A prototype circuit for energy recovery demonstrated that the maximum count rates can be increased by more than 10 times comparing to the pulse-shaping method, and the energy resolution is as good as pulse shaping (or fixed integration) at low count rates. At 2 x 10 6 events/sec on NaI(Tl), the true counts acquired with this method is 3.3 times more than the delay-line clipping method (256 ns clipping) due to events recovered from pileups. Pulse-height spectra up to 3.5 x 10 6 events/sec have been studied. Monte Carlo simulation studies have been performed for image-quality comparisons between different processing methods

  5. Implementation of a nonlinear filter for online nuclear counting

    International Nuclear Information System (INIS)

    Coulon, R.; Dumazert, J.; Kondrasovs, V.; Normand, S.

    2016-01-01

    Nuclear counting is a challenging task for nuclear instrumentation because of the stochastic nature of radioactivity. Event counting has to be processed and filtered to determine a stable count rate value and perform variation monitoring of the measured event. An innovative approach for nuclear counting is presented in this study, improving response time and maintaining count rate stability. Some nonlinear filters providing a local maximum likelihood estimation of the signal have been recently developed, which have been tested and compared with conventional linear filters. A nonlinear filter thus developed shows significant performance in terms of response time and measurement precision. The filter also presents the specificity of easy embedment into digital signal processor (DSP) electronics based on field-programmable gate arrays (FPGA) or microcontrollers, compatible with real-time requirements. © 2001 Elsevier Science. All rights reserved. - Highlights: • An efficient approach based on nonlinear filtering has been implemented. • The hypothesis test provides a local maximum likelihood estimation of the count rate. • The filter ensures an optimal compromise between precision and response time.

  6. Advanced digital counting rate meter for gamma ray logging

    International Nuclear Information System (INIS)

    Kannan, S.; Meenakshi Sundari, A.; Rai, A.K.

    2013-01-01

    A compact, hand held controller based Advanced Digital Count Rate Meter (ADCRM) as a replacement of bulky Count Rate (analog) Meters (SBL-2A) was designed, developed and fabricated to carry out Gamma-Ray borehole logging with Geiger Muller (GM) tubes and Scintillation (SC) detectors. In the hardware the functionality of analog meter simulation, digital counting of gamma events and auto reference adjustment to use different length of armour cable winches were implemented. The in-built software evaluates grade in ppm and at the end of logging, the reports are prepared automatically. ADCRM was developed in-house to assist the uranium mineral exploration in AMD. (author)

  7. A burst-mode photon counting receiver with automatic channel estimation and bit rate detection

    Science.gov (United States)

    Rao, Hemonth G.; DeVoe, Catherine E.; Fletcher, Andrew S.; Gaschits, Igor D.; Hakimi, Farhad; Hamilton, Scott A.; Hardy, Nicholas D.; Ingwersen, John G.; Kaminsky, Richard D.; Moores, John D.; Scheinbart, Marvin S.; Yarnall, Timothy M.

    2016-04-01

    We demonstrate a multi-rate burst-mode photon-counting receiver for undersea communication at data rates up to 10.416 Mb/s over a 30-foot water channel. To the best of our knowledge, this is the first demonstration of burst-mode photon-counting communication. With added attenuation, the maximum link loss is 97.1 dB at λ=517 nm. In clear ocean water, this equates to link distances up to 148 meters. For λ=470 nm, the achievable link distance in clear ocean water is 450 meters. The receiver incorporates soft-decision forward error correction (FEC) based on a product code of an inner LDPC code and an outer BCH code. The FEC supports multiple code rates to achieve error-free performance. We have selected a burst-mode receiver architecture to provide robust performance with respect to unpredictable channel obstructions. The receiver is capable of on-the-fly data rate detection and adapts to changing levels of signal and background light. The receiver updates its phase alignment and channel estimates every 1.6 ms, allowing for rapid changes in water quality as well as motion between transmitter and receiver. We demonstrate on-the-fly rate detection, channel BER within 0.2 dB of theory across all data rates, and error-free performance within 1.82 dB of soft-decision capacity across all tested code rates. All signal processing is done in FPGAs and runs continuously in real time.

  8. General Theory of Decoy-State Quantum Cryptography with Dark Count Rate Fluctuation

    International Nuclear Information System (INIS)

    Xiang, Gao; Shi-Hai, Sun; Lin-Mei, Liang

    2009-01-01

    The existing theory of decoy-state quantum cryptography assumes that the dark count rate is a constant, but in practice there exists fluctuation. We develop a new scheme of the decoy state, achieve a more practical key generation rate in the presence of fluctuation of the dark count rate, and compare the result with the result of the decoy-state without fluctuation. It is found that the key generation rate and maximal secure distance will be decreased under the influence of the fluctuation of the dark count rate

  9. Bayesian analysis of energy and count rate data for detection of low count rate radioactive sources

    Energy Technology Data Exchange (ETDEWEB)

    Klumpp, John [Colorado State University, Department of Environmental and Radiological Health Sciences, Molecular and Radiological Biosciences Building, Colorado State University, Fort Collins, Colorado, 80523 (United States)

    2013-07-01

    We propose a radiation detection system which generates its own discrete sampling distribution based on past measurements of background. The advantage to this approach is that it can take into account variations in background with respect to time, location, energy spectra, detector-specific characteristics (i.e. different efficiencies at different count rates and energies), etc. This would therefore be a 'machine learning' approach, in which the algorithm updates and improves its characterization of background over time. The system would have a 'learning mode,' in which it measures and analyzes background count rates, and a 'detection mode,' in which it compares measurements from an unknown source against its unique background distribution. By characterizing and accounting for variations in the background, general purpose radiation detectors can be improved with little or no increase in cost. The statistical and computational techniques to perform this kind of analysis have already been developed. The necessary signal analysis can be accomplished using existing Bayesian algorithms which account for multiple channels, multiple detectors, and multiple time intervals. Furthermore, Bayesian machine-learning techniques have already been developed which, with trivial modifications, can generate appropriate decision thresholds based on the comparison of new measurements against a nonparametric sampling distribution. (authors)

  10. Exercise-induced maximum metabolic rate scaled to body mass by ...

    African Journals Online (AJOL)

    Exercise-induced maximum metabolic rate scaled to body mass by the fractal ... rate scaling is that exercise-induced maximum aerobic metabolic rate (MMR) is ... muscle stress limitation, and maximized oxygen delivery and metabolic rates.

  11. Smart pile-up consideration for evaluation of high count rate EDS spectra

    International Nuclear Information System (INIS)

    Eggert, F; Anderhalt, R; Nicolosi, J; Elam, T

    2012-01-01

    This work describes a new pile-up consideration for the very high count rate spectra which are possible to acquire with silicon drift detector (SDD) technology. Pile-up effects are the major and still remaining challenge with the use of SDD for EDS in scanning electron microscopes (SEM) with ultra thin windows for soft X-ray detection. The ability to increase the count rates up to a factor of 100 compared with conventional Si(Li) detectors, comes with the problem that the pile-up recognition (pile-up rejection) in pulse processors is not able to improve by the same order of magnitude, just only with a factor of about 3. Therefore, it is common that spectra will show significant pile-up effects if count rates of more than 10000 counts per second (10 kcps) are used. These false counts affect both automatic qualitative analysis and quantitative evaluation of the spectra. The new idea is to use additional inputs for pile-up calculation to shift the applicability towards very high count rates of up to 200 kcps and more, which can be easily acquired with the SDD. The additional input is the 'known' (estimated) background distribution, calculated iteratively during all automated qualitative or quantitative evaluations. This additional knowledge gives the opportunity for self adjustment of the pile-up calculation parameters and avoids over-corrections which challenge the evaluation as well as the pile-up artefacts themselves. With the proposed method the pile-up correction is no longer a 'correction' but an integral part of all spectra evaluation steps. Examples for the application are given with evaluation of very high count rate spectra.

  12. Studies on the Pulse Rate, Pedometer Count and Satisfactoin Degree at Various Exercise

    OpenAIRE

    小原, 史朗

    2004-01-01

    This investigation examined whether free exercise of students became good stimulation of breathing circulation function from relation of pulse rate and pedometer count. And, I examined it on satisfaction degree after exercise. Object person was 432 man students (total of 1391) and 94 woman students (total of 472). As for relation of pulse rate and pedometer count, statistical meaning was recognized by man and women. The exercise that a pulse rate and pedometer count were high together seemed ...

  13. Study on the behaviour of timing photomultipliers at a high counting rate

    International Nuclear Information System (INIS)

    Gladyshev, D.A.; Li, B.N.; Yunusov, Kh.R.

    1978-01-01

    Variations in the amplification factor K of a photomultiplier (PMU) with the accuracy of 1% in a pulse mode are studied. Measurements were performed by means of a light pulse generator based on a light diode which generates pulses at the repetition rate of 250-10 5 pulse/s. Relative variations in K were determined by the position of the peak gravity centre from the light diode using a pulse analyzer and a frequency meter. Results of PM testing show that, at a sudden counting rate increase, the amplification increases during the time period less than, the measurement time (less than 1 s) and returns to the stationary value. When the counting rate returns from 10 5 pulse/s to the initial value of 250 pulse/s, the amplification decreases and than increases to stationary value. The total time of K stabilization after counting rate applying constitutes 10-70 min. Restoration of K after counting rate removal occurs to be much slower, during 3 hr. 40 min. K values varied from 1 to 12%

  14. Testing the count rate performance of the scintillation camera by exponential attenuation: Decaying source; Multiple filters

    International Nuclear Information System (INIS)

    Adams, R.; Mena, I.

    1988-01-01

    An algorithm and two fortrAN programs have been developed to evaluate the count rate performance of scintillation cameras from count rates reduced exponentially, either by a decaying source or by filtration. The first method is used with short-lived radionuclides such as 191 /sup m/Ir or 191 /sup m/Au. The second implements a National Electrical Manufacturers' Association (NEMA) protocol in which the count rate from a source of 191 /sup m/Tc is attenuated by a varying number of copper filters stacked over it. The count rate at each data point is corrected for deadtime loss after assigning an arbitrary deadtime (tau). A second-order polynomial equation is fitted to the logarithms of net count rate values: ln(R) = A+BT+CT 2 where R is the net corrected count rate (cps), and T is the elapsed time (or the filter thickness in the NEMA method). Depending on C, tau is incremented or decremented iteratively, and the count rate corrections and curve fittings are repeated until C approaches zero, indicating a correct value of the deadtime (tau). The program then plots the measured count rate versus the corrected count rate values

  15. Accuracy in activation analysis: count rate effects

    International Nuclear Information System (INIS)

    Lindstrom, R.M.; Fleming, R.F.

    1980-01-01

    The accuracy inherent in activation analysis is ultimately limited by the uncertainty of counting statistics. When careful attention is paid to detail, several workers have shown that all systematic errors can be reduced to an insignificant fraction of the total uncertainty, even when the statistical limit is well below one percent. A matter of particular importance is the reduction of errors due to high counting rate. The loss of counts due to random coincidence (pulse pileup) in the amplifier and to digitization time in the ADC may be treated as a series combination of extending and non-extending dead times, respectively. The two effects are experimentally distinct. Live timer circuits in commercial multi-channel analyzers compensate properly for ADC dead time for long-lived sources, but not for pileup. Several satisfactory solutions are available, including pileup rejection and dead time correction circuits, loss-free ADCs, and computed corrections in a calibrated system. These methods are sufficiently reliable and well understood that a decaying source can be measured routinely with acceptably small errors at a dead time as high as 20 percent

  16. Measures to handle accidental contamination of persons with standard counting devices available in a department of nuclear medicine

    International Nuclear Information System (INIS)

    Aiginger, H.; Lauer, D.; Unfried, E.; Koenig, F.; Ogris, E.

    1998-01-01

    The assets and shortcomings of a well-type NaI detector and a Ge detector were examined using the Marinelli geometry, test tube geometry, and beaker geometry. Plots of the efficiency vs. energy (efficiency calibration), recorded time vs. true time (dead-time effects), and maximum activity vs. energy are reproduced. A high counting efficiency is typical of the scintillation detector in the well for the test tube geometry, particularly in the low energy range. For energies higher than 100 keV, the counting efficiency decreases because of the increasing penetration of the detector bulk by high-energy photons. For the germanium detector, the highest counting efficiency was achieved in the beaker geometry. A linear relationship exists between the calculated and measured counts at the beginning of the recorded curve for both systems. For the well-type detector the maximum detectable count rate was about 30 kcps, the linearity of the plot of the true count rate was guaranteed up to 10 kcps. Dead time correction was to be made at higher count rates. For the germanium detector the maximum detectable count rate was only about 8 kcps due to the longer dead time, the linear segment, however, was longer than for the scintillation detector. It is concluded that although the maximum detectable count rate of the germanium detector is lower, higher true activities can be detected with it owing to the lower detection efficiency. The well-type scintillation detector is advantageous for the test tube geometry. (P.A.)

  17. 5 CFR 531.221 - Maximum payable rate rule.

    Science.gov (United States)

    2010-01-01

    ... before the reassignment. (ii) If the rate resulting from the geographic conversion under paragraph (c)(2... previous rate (i.e., the former special rate after the geographic conversion) with the rates on the current... 5 Administrative Personnel 1 2010-01-01 2010-01-01 false Maximum payable rate rule. 531.221...

  18. High rate 4π β-γ coincidence counting system

    International Nuclear Information System (INIS)

    Johnson, L.O.; Gehrke, R.J.

    1978-01-01

    A high count rate 4π β-γ coincidence counting system for the determination of absolute disintegration rates of short half-life radionuclides is described. With this system the dead time per pulse is minimized by not stretching any pulses beyond the width necessary to satisfy overlap coincidence requirements. The equations used to correct for the β, γ, and coincidence channel dead times and for accidental coincidences are presented but not rigorously developed. Experimental results are presented for a decaying source of 56 Mn initially at 2 x 10 6 d/s and a set of 60 Co sources of accurately known source strengths varying from 10 3 to 2 x 10 6 d/s. A check of the accidental coincidence equation for the case of two independent sources with varying source strengths is presented

  19. Fluctuations and dark count rates in superconducting NbN single-photon detectors

    International Nuclear Information System (INIS)

    Engel, Andreas; Semenov, Alexei; Huebers, Heinz-Wilhelm; Il'in, Kostya; Siegel, Michael

    2005-01-01

    We measured the temperature- and current-dependence of dark count rates of a superconducting singlephoton detector. The detector's key element is a 84 nm wide meander strip line fabricated from a 5 nm thick NbN film. Due to its reduced dimensions various types of fluctuations can cause temporal and localized transitions into a resistive state leading to dark count events. Adopting a recent refinement of the hotspot model we achieve a satisfying description of the experimental dark count rates taking into account fluctuations of the Cooper-pair density and current-assisted unbinding of vortex-antivortex pairs. (copyright 2005 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  20. Reducing the Teen Death Rate. KIDS COUNT Indicator Brief

    Science.gov (United States)

    Shore, Rima; Shore, Barbara

    2009-01-01

    Life continues to hold considerable risk for adolescents in the United States. In 2006, the teen death rate stood at 64 deaths per 100,000 teens (13,739 teens) (KIDS COUNT Data Center, 2009). Although it has declined by 4 percent since 2000, the rate of teen death in this country remains substantially higher than in many peer nations, based…

  1. Method and system of simulating nuclear power plant count rate for training purposes

    International Nuclear Information System (INIS)

    Alliston, W.H.; Koenig, R.H.

    1975-01-01

    A method and system are described for the real-time simulation of the dynamic operation of a nuclear power plant in which nuclear flux rate counters are provided for monitoring the rate of nuclear fission of the reactor. The system utilizes apparatus that includes digital computer means for calculating data relating to the rate of nuclear fission of a simulated reactor model, which rate is controlled in accordance with the operation of control panel devices. A digital number from the computer corresponding to the flux rate controls an oscillator driven counter means to produce a pulse after a predetermined count. This pulse controls an oscillator driven polynomial counter to count a random number that controls a third counter in accordance with pulse from the first counter to produce a random fission count for operating the meters. (U.S.)

  2. Dinosaur Metabolism and the Allometry of Maximum Growth Rate

    OpenAIRE

    Myhrvold, Nathan P.

    2016-01-01

    The allometry of maximum somatic growth rate has been used in prior studies to classify the metabolic state of both extant vertebrates and dinosaurs. The most recent such studies are reviewed, and their data is reanalyzed. The results of allometric regressions on growth rate are shown to depend on the choice of independent variable; the typical choice used in prior studies introduces a geometric shear transformation that exaggerates the statistical power of the regressions. The maximum growth...

  3. Accurate disintegration-rate measurement of 55Fe by liquid scintillation counting

    International Nuclear Information System (INIS)

    Steyn, J.; Oberholzer, P.; Botha, S.M.

    1979-01-01

    A method involving liquid scintillation counting is described for the accurate measurement of disintegration rate of 55 Fe. The method is based on the use of calculated efficiency functions together with either of the nuclides 54 Mn and 51 Cr as internal standards for measurement of counting efficiencies by coincidence counting. The method was used by the NAC during a recent international intercomparison of radioactivity measurements, and a summary of the results obtained by nine participating laboratories is presented. A spread in results of several percent is evident [af

  4. 44 CFR 208.12 - Maximum Pay Rate Table.

    Science.gov (United States)

    2010-10-01

    ...) Physicians. DHS uses the latest Special Salary Rate Table Number 0290 for Medical Officers (Clinical... Personnel, in which case the Maximum Pay Rate Table would not apply. (3) Compensation for Sponsoring Agency... organizations, e.g., HMOs or medical or engineering professional associations, under the revised definition of...

  5. Soft X ray spectrometry at high count rates

    International Nuclear Information System (INIS)

    Blanc, P.; Brouquet, P.; Uhre, N.

    1978-06-01

    Two modifications of the classical method of X-ray spectrometry by a semi-conductor diode permit a count rate of 10 5 c/s with an energy resolution of 350 eV. With a specially constructed pulse height analyzer, this detector can measure four spectra of 5 ms each, in the range of 1-30 keV, during a plasma shot

  6. Maximum organic carbon limits at different melter feed rates (U)

    International Nuclear Information System (INIS)

    Choi, A.S.

    1995-01-01

    This report documents the results of a study to assess the impact of varying melter feed rates on the maximum total organic carbon (TOC) limits allowable in the DWPF melter feed. Topics discussed include: carbon content; feed rate; feed composition; melter vapor space temperature; combustion and dilution air; off-gas surges; earlier work on maximum TOC; overview of models; and the results of the work completed

  7. Platelet Counts, MPV and PDW in Culture Proven and Probable Neonatal Sepsis and Association of Platelet Counts with Mortality Rate

    International Nuclear Information System (INIS)

    Ahmad, M. S.; Waheed, A.

    2014-01-01

    Objective: To determine frequency of thrombocytopenia and thrombocytosis, the MPV (mean platelet volume) and PDW (platelet distribution width) in patients with probable and culture proven neonatal sepsis and determine any association between platelet counts and mortality rate. Study Design: Descriptive analytical study. Place and Duration of Study: NICU, Fazle Omar Hospital, from January 2011 to December 2012. Methodology: Cases of culture proven and probable neonatal sepsis, admitted in Fazle Omar Hospital, Rabwah, were included in the study. Platelet counts, MPV and PDW of the cases were recorded. Mortality was documented. Frequencies of thrombocytopenia ( 450000/mm3) were ascertained. Mortality rates in different groups according to platelet counts were calculated and compared by chi-square test to check association. Results: Four hundred and sixty nine patients were included; 68 (14.5%) of them died. One hundred and thirty six (29%) had culture proven sepsis, and 333 (71%) were categorized as probable sepsis. Thrombocytopenia was present in 116 (24.7%), and thrombocytosis was present in 36 (7.7%) cases. Median platelet count was 213.0/mm3. Twenty eight (27.7%) patients with thrombocytopenia, and 40 (12.1%) cases with normal or raised platelet counts died (p < 0.001). Median MPV was 9.30, and median PDW was 12.30. MPV and PDW of the patients who died and who were discharged were not significantly different from each other. Conclusion: Thrombocytopenia is a common complication of neonatal sepsis. Those with thrombocytopenia have higher mortality rate. No significant difference was present between PDW and MPV of the cases who survived and died. (author)

  8. Dinosaur Metabolism and the Allometry of Maximum Growth Rate.

    Science.gov (United States)

    Myhrvold, Nathan P

    2016-01-01

    The allometry of maximum somatic growth rate has been used in prior studies to classify the metabolic state of both extant vertebrates and dinosaurs. The most recent such studies are reviewed, and their data is reanalyzed. The results of allometric regressions on growth rate are shown to depend on the choice of independent variable; the typical choice used in prior studies introduces a geometric shear transformation that exaggerates the statistical power of the regressions. The maximum growth rates of extant groups are found to have a great deal of overlap, including between groups with endothermic and ectothermic metabolism. Dinosaur growth rates show similar overlap, matching the rates found for mammals, reptiles and fish. The allometric scaling of growth rate with mass is found to have curvature (on a log-log scale) for many groups, contradicting the prevailing view that growth rate allometry follows a simple power law. Reanalysis shows that no correlation between growth rate and basal metabolic rate (BMR) has been demonstrated. These findings drive a conclusion that growth rate allometry studies to date cannot be used to determine dinosaur metabolism as has been previously argued.

  9. Dinosaur Metabolism and the Allometry of Maximum Growth Rate

    Science.gov (United States)

    Myhrvold, Nathan P.

    2016-01-01

    The allometry of maximum somatic growth rate has been used in prior studies to classify the metabolic state of both extant vertebrates and dinosaurs. The most recent such studies are reviewed, and their data is reanalyzed. The results of allometric regressions on growth rate are shown to depend on the choice of independent variable; the typical choice used in prior studies introduces a geometric shear transformation that exaggerates the statistical power of the regressions. The maximum growth rates of extant groups are found to have a great deal of overlap, including between groups with endothermic and ectothermic metabolism. Dinosaur growth rates show similar overlap, matching the rates found for mammals, reptiles and fish. The allometric scaling of growth rate with mass is found to have curvature (on a log-log scale) for many groups, contradicting the prevailing view that growth rate allometry follows a simple power law. Reanalysis shows that no correlation between growth rate and basal metabolic rate (BMR) has been demonstrated. These findings drive a conclusion that growth rate allometry studies to date cannot be used to determine dinosaur metabolism as has been previously argued. PMID:27828977

  10. Simple bounds for counting processes with monotone rate of occurrence of failures

    International Nuclear Information System (INIS)

    Kaminskiy, Mark P.

    2007-01-01

    The article discusses some aspects of analogy between certain classes of distributions used as models for time to failure of nonrepairable objects, and the counting processes used as models for failure process for repairable objects. The notion of quantiles for the counting processes with strictly increasing cumulative intensity function is introduced. The classes of counting processes with increasing (decreasing) rate of occurrence of failures are considered. For these classes, the useful nonparametric bounds for cumulative intensity function based on one known quantile are obtained. These bounds, which can be used for repairable objects, are similar to the bounds introduced by Barlow and Marshall [Barlow, R. Marshall, A. Bounds for distributions with monotone hazard rate, I and II. Ann Math Stat 1964; 35: 1234-74] for IFRA (DFRA) time to failure distributions applicable to nonrepairable objects

  11. Relationship between salivary flow rates and Candida counts in subjects with xerostomia.

    Science.gov (United States)

    Torres, Sandra R; Peixoto, Camila Bernardo; Caldas, Daniele Manhães; Silva, Eline Barboza; Akiti, Tiyomi; Nucci, Márcio; de Uzeda, Milton

    2002-02-01

    This study evaluated the relationship between salivary flow and Candida colony counts in the saliva of patients with xerostomia. Sialometry and Candida colony-forming unit (CFU) counts were taken from 112 subjects who reported xerostomia in a questionnaire. Chewing-stimulated whole saliva was collected and streaked in Candida plates and counted in 72 hours. Species identification was accomplished under standard methods. There was a significant inverse relationship between salivary flow and Candida CFU counts (P =.007) when subjects with high colony counts were analyzed (cutoff point of 400 or greater CFU/mL). In addition, the median sialometry of men was significantly greater than that of women (P =.003), even after controlling for confounding variables like underlying disease and medications. Sjögren's syndrome was associated with low salivary flow rate (P =.007). There was no relationship between the median Candida CFU counts and gender or age. There was a high frequency (28%) of mixed colonization. Candida albicans was the most frequent species, followed by C parapsilosis, C tropicalis, and C krusei. In subjects with high Candida CFU counts there was an inverse relationship between salivary flow and Candida CFU counts.

  12. 47 CFR 1.1507 - Rulemaking on maximum rates for attorney fees.

    Science.gov (United States)

    2010-10-01

    ... 47 Telecommunication 1 2010-10-01 2010-10-01 false Rulemaking on maximum rates for attorney fees... § 1.1507 Rulemaking on maximum rates for attorney fees. (a) If warranted by an increase in the cost of... types of proceedings), the Commission may adopt regulations providing that attorney fees may be awarded...

  13. BMRC: A Bitmap-Based Maximum Range Counting Approach for Temporal Data in Sensor Monitoring Networks

    Directory of Open Access Journals (Sweden)

    Bin Cao

    2017-09-01

    Full Text Available Due to the rapid development of the Internet of Things (IoT, many feasible deployments of sensor monitoring networks have been made to capture the events in physical world, such as human diseases, weather disasters and traffic accidents, which generate large-scale temporal data. Generally, the certain time interval that results in the highest incidence of a severe event has significance for society. For example, there exists an interval that covers the maximum number of people who have the same unusual symptoms, and knowing this interval can help doctors to locate the reason behind this phenomenon. As far as we know, there is no approach available for solving this problem efficiently. In this paper, we propose the Bitmap-based Maximum Range Counting (BMRC approach for temporal data generated in sensor monitoring networks. Since sensor nodes can update their temporal data at high frequency, we present a scalable strategy to support the real-time insert and delete operations. The experimental results show that the BMRC outperforms the baseline algorithm in terms of efficiency.

  14. On the fast response of charnel electron multipliers in coUnting mode operation

    International Nuclear Information System (INIS)

    Belyaevskij, O.A.; Gladyshev, I.L.; Korobochko, Yu.S.; Mineev, V.I.

    1983-01-01

    Dependences of amplitude distribution of pulses at the outlet of channel electron multipliers (CEM) and effectiveness of monitoring on counting rate at different supply voltages are determined. It is shown that the maximUm counting rate of CEM runs into 6x10 5 s -1 at short-term and 10 5 s -1 at long-term operation using monitoring eqUipment with operation threshold of 2.5 mV

  15. A pulse stacking method of particle counting applied to position sensitive detection

    International Nuclear Information System (INIS)

    Basilier, E.

    1976-03-01

    A position sensitive particle counting system is described. A cyclic readout imaging device serves as an intermediate information buffer. Pulses are allowed to stack in the imager at very high counting rates. Imager noise is completely discriminated to provide very wide dynamic range. The system has been applied to a detector using cascaded microchannel plates. Pulse height spread produced by the plates causes some loss of information. The loss is comparable to the input loss of the plates. The improvement in maximum counting rate is several hundred times over previous systems that do not permit pulse stacking. (Auth.)

  16. Allometries of Maximum Growth Rate versus Body Mass at Maximum Growth Indicate That Non-Avian Dinosaurs Had Growth Rates Typical of Fast Growing Ectothermic Sauropsids

    Science.gov (United States)

    Werner, Jan; Griebeler, Eva Maria

    2014-01-01

    We tested if growth rates of recent taxa are unequivocally separated between endotherms and ectotherms, and compared these to dinosaurian growth rates. We therefore performed linear regression analyses on the log-transformed maximum growth rate against log-transformed body mass at maximum growth for extant altricial birds, precocial birds, eutherians, marsupials, reptiles, fishes and dinosaurs. Regression models of precocial birds (and fishes) strongly differed from Case’s study (1978), which is often used to compare dinosaurian growth rates to those of extant vertebrates. For all taxonomic groups, the slope of 0.75 expected from the Metabolic Theory of Ecology was statistically supported. To compare growth rates between taxonomic groups we therefore used regressions with this fixed slope and group-specific intercepts. On average, maximum growth rates of ectotherms were about 10 (reptiles) to 20 (fishes) times (in comparison to mammals) or even 45 (reptiles) to 100 (fishes) times (in comparison to birds) lower than in endotherms. While on average all taxa were clearly separated from each other, individual growth rates overlapped between several taxa and even between endotherms and ectotherms. Dinosaurs had growth rates intermediate between similar sized/scaled-up reptiles and mammals, but a much lower rate than scaled-up birds. All dinosaurian growth rates were within the range of extant reptiles and mammals, and were lower than those of birds. Under the assumption that growth rate and metabolic rate are indeed linked, our results suggest two alternative interpretations. Compared to other sauropsids, the growth rates of studied dinosaurs clearly indicate that they had an ectothermic rather than an endothermic metabolic rate. Compared to other vertebrate growth rates, the overall high variability in growth rates of extant groups and the high overlap between individual growth rates of endothermic and ectothermic extant species make it impossible to rule out either

  17. Allometries of maximum growth rate versus body mass at maximum growth indicate that non-avian dinosaurs had growth rates typical of fast growing ectothermic sauropsids.

    Science.gov (United States)

    Werner, Jan; Griebeler, Eva Maria

    2014-01-01

    We tested if growth rates of recent taxa are unequivocally separated between endotherms and ectotherms, and compared these to dinosaurian growth rates. We therefore performed linear regression analyses on the log-transformed maximum growth rate against log-transformed body mass at maximum growth for extant altricial birds, precocial birds, eutherians, marsupials, reptiles, fishes and dinosaurs. Regression models of precocial birds (and fishes) strongly differed from Case's study (1978), which is often used to compare dinosaurian growth rates to those of extant vertebrates. For all taxonomic groups, the slope of 0.75 expected from the Metabolic Theory of Ecology was statistically supported. To compare growth rates between taxonomic groups we therefore used regressions with this fixed slope and group-specific intercepts. On average, maximum growth rates of ectotherms were about 10 (reptiles) to 20 (fishes) times (in comparison to mammals) or even 45 (reptiles) to 100 (fishes) times (in comparison to birds) lower than in endotherms. While on average all taxa were clearly separated from each other, individual growth rates overlapped between several taxa and even between endotherms and ectotherms. Dinosaurs had growth rates intermediate between similar sized/scaled-up reptiles and mammals, but a much lower rate than scaled-up birds. All dinosaurian growth rates were within the range of extant reptiles and mammals, and were lower than those of birds. Under the assumption that growth rate and metabolic rate are indeed linked, our results suggest two alternative interpretations. Compared to other sauropsids, the growth rates of studied dinosaurs clearly indicate that they had an ectothermic rather than an endothermic metabolic rate. Compared to other vertebrate growth rates, the overall high variability in growth rates of extant groups and the high overlap between individual growth rates of endothermic and ectothermic extant species make it impossible to rule out either of

  18. Allometries of maximum growth rate versus body mass at maximum growth indicate that non-avian dinosaurs had growth rates typical of fast growing ectothermic sauropsids.

    Directory of Open Access Journals (Sweden)

    Jan Werner

    Full Text Available We tested if growth rates of recent taxa are unequivocally separated between endotherms and ectotherms, and compared these to dinosaurian growth rates. We therefore performed linear regression analyses on the log-transformed maximum growth rate against log-transformed body mass at maximum growth for extant altricial birds, precocial birds, eutherians, marsupials, reptiles, fishes and dinosaurs. Regression models of precocial birds (and fishes strongly differed from Case's study (1978, which is often used to compare dinosaurian growth rates to those of extant vertebrates. For all taxonomic groups, the slope of 0.75 expected from the Metabolic Theory of Ecology was statistically supported. To compare growth rates between taxonomic groups we therefore used regressions with this fixed slope and group-specific intercepts. On average, maximum growth rates of ectotherms were about 10 (reptiles to 20 (fishes times (in comparison to mammals or even 45 (reptiles to 100 (fishes times (in comparison to birds lower than in endotherms. While on average all taxa were clearly separated from each other, individual growth rates overlapped between several taxa and even between endotherms and ectotherms. Dinosaurs had growth rates intermediate between similar sized/scaled-up reptiles and mammals, but a much lower rate than scaled-up birds. All dinosaurian growth rates were within the range of extant reptiles and mammals, and were lower than those of birds. Under the assumption that growth rate and metabolic rate are indeed linked, our results suggest two alternative interpretations. Compared to other sauropsids, the growth rates of studied dinosaurs clearly indicate that they had an ectothermic rather than an endothermic metabolic rate. Compared to other vertebrate growth rates, the overall high variability in growth rates of extant groups and the high overlap between individual growth rates of endothermic and ectothermic extant species make it impossible to rule

  19. Maximum Entropy and Theory Construction: A Reply to Favretti

    Directory of Open Access Journals (Sweden)

    John Harte

    2018-04-01

    Full Text Available In the maximum entropy theory of ecology (METE, the form of a function describing the distribution of abundances over species and metabolic rates over individuals in an ecosystem is inferred using the maximum entropy inference procedure. Favretti shows that an alternative maximum entropy model exists that assumes the same prior knowledge and makes predictions that differ from METE’s. He shows that both cannot be correct and asserts that his is the correct one because it can be derived from a classic microstate-counting calculation. I clarify here exactly what the core entities and definitions are for METE, and discuss the relevance of two critical issues raised by Favretti: the existence of a counting procedure for microstates and the choices of definition of the core elements of a theory. I emphasize that a theorist controls how the core entities of his or her theory are defined, and that nature is the final arbiter of the validity of a theory.

  20. Detectability of changes in cosmic-ray counting rate measured with the Liulin detector

    International Nuclear Information System (INIS)

    Malusek, A.; Kubancak, J.; Ambrozova, I.

    2011-05-01

    Experimental data are needed to improve and validate models predicting the dynamics of solar particle events because the mechanisms of processes leading to the acceleration of solar energetic particles are not yet fully understood. The aim of this work was to examine whether the spectrometer of deposited energy, Liulin, positioned at the Lomnický štít mountain observatory can collect such data. Decision thresholds and detection limits for the increase or decrease in the average number of particles detected by Liulin were determined for a period in February 2011. The changes in counts corresponding to the decision thresholds and detection limits relative to the average number of detected particles were about 17% and 33%, respectively. The Forbush decrease with a maximum change of about 6.8%, which started on February 18, was detectable neither during the 10-minute acquisition time nor during any other, longer period. Statistical analysis showed that an acquisition time about 7 hours would be needed to detect a 5% decrease. As this time was shorter than the duration of the Forbush decrease (about 56 hours), we theorize that the current placement of the Liulin detector inside a living room shielded by a thick concrete ceiling may have had an adverse impact on the detectability of the the cosmic ray counting rate decrease. To test this hypothesis, we recommend positioning the Liulin detector outside the main observatory building.. (author)

  1. On-line statistical processing of radiation detector pulse trains with time-varying count rates

    International Nuclear Information System (INIS)

    Apostolopoulos, G.

    2008-01-01

    Statistical analysis is of primary importance for the correct interpretation of nuclear measurements, due to the inherent random nature of radioactive decay processes. This paper discusses the application of statistical signal processing techniques to the random pulse trains generated by radiation detectors. The aims of the presented algorithms are: (i) continuous, on-line estimation of the underlying time-varying count rate θ(t) and its first-order derivative dθ/dt; (ii) detection of abrupt changes in both of these quantities and estimation of their new value after the change point. Maximum-likelihood techniques, based on the Poisson probability distribution, are employed for the on-line estimation of θ and dθ/dt. Detection of abrupt changes is achieved on the basis of the generalized likelihood ratio statistical test. The properties of the proposed algorithms are evaluated by extensive simulations and possible applications for on-line radiation monitoring are discussed

  2. Ballistic deficit correction methods for large Ge detectors-high counting rate study

    International Nuclear Information System (INIS)

    Duchene, G.; Moszynski, M.

    1995-01-01

    This study presents different ballistic correction methods versus input count rate (from 3 to 50 kcounts/s) using four large Ge detectors of about 70 % relative efficiency. It turns out that the Tennelec TC245 linear amplifier in the BDC mode (Hinshaw method) is the best compromise for energy resolution throughout. All correction methods lead to narrow sum-peaks indistinguishable from single Γ lines. The full energy peak throughput is found representative of the pile-up inspection dead time of the corrector circuits. This work also presents a new and simple representation, plotting simultaneously energy resolution and throughput versus input count rate. (TEC). 12 refs., 11 figs

  3. ChromAIX2: A large area, high count-rate energy-resolving photon counting ASIC for a Spectral CT Prototype

    Science.gov (United States)

    Steadman, Roger; Herrmann, Christoph; Livne, Amir

    2017-08-01

    Spectral CT based on energy-resolving photon counting detectors is expected to deliver additional diagnostic value at a lower dose than current state-of-the-art CT [1]. The capability of simultaneously providing a number of spectrally distinct measurements not only allows distinguishing between photo-electric and Compton interactions but also discriminating contrast agents that exhibit a K-edge discontinuity in the absorption spectrum, referred to as K-edge Imaging [2]. Such detectors are based on direct converting sensors (e.g. CdTe or CdZnTe) and high-rate photon counting electronics. To support the development of Spectral CT and show the feasibility of obtaining rates exceeding 10 Mcps/pixel (Poissonian observed count-rate), the ChromAIX ASIC has been previously reported showing 13.5 Mcps/pixel (150 Mcps/mm2 incident) [3]. The ChromAIX has been improved to offer the possibility of a large area coverage detector, and increased overall performance. The new ASIC is called ChromAIX2, and delivers count-rates exceeding 15 Mcps/pixel with an rms-noise performance of approximately 260 e-. It has an isotropic pixel pitch of 500 μm in an array of 22×32 pixels and is tile-able on three of its sides. The pixel topology consists of a two stage amplifier (CSA and Shaper) and a number of test features allowing to thoroughly characterize the ASIC without a sensor. A total of 5 independent thresholds are also available within each pixel, allowing to acquire 5 spectrally distinct measurements simultaneously. The ASIC also incorporates a baseline restorer to eliminate excess currents induced by the sensor (e.g. dark current and low frequency drifts) which would otherwise cause an energy estimation error. In this paper we report on the inherent electrical performance of the ChromAXI2 as well as measurements obtained with CZT (CdZnTe)/CdTe sensors and X-rays and radioactive sources.

  4. Gain reduction due to space charge at high counting rates in multiwire proportional chambers

    International Nuclear Information System (INIS)

    Smith, G.C.; Mathieson, E.

    1986-10-01

    Measurements with a small MWPC of gas gain reduction, due to ion space charge at high counting rates, have been compared with theoretical predictions. The quantity ln(q/q 0 )/(q/q 0 ), where (q/q 0 ) is the relative reduced avalanche charge, has been found to be closely proportional to count rate, as predicted. The constant of proportionality is in good agreement with calculations made with a modified version of the original, simplified theory

  5. 5 CFR 9901.312 - Maximum rates of base salary and adjusted salary.

    Science.gov (United States)

    2010-01-01

    ... 5 Administrative Personnel 3 2010-01-01 2010-01-01 false Maximum rates of base salary and adjusted salary. 9901.312 Section 9901.312 Administrative Personnel DEPARTMENT OF DEFENSE HUMAN RESOURCES....312 Maximum rates of base salary and adjusted salary. (a) Subject to § 9901.105, the Secretary may...

  6. The scaling of maximum and basal metabolic rates of mammals and birds

    Science.gov (United States)

    Barbosa, Lauro A.; Garcia, Guilherme J. M.; da Silva, Jafferson K. L.

    2006-01-01

    Allometric scaling is one of the most pervasive laws in biology. Its origin, however, is still a matter of dispute. Recent studies have established that maximum metabolic rate scales with an exponent larger than that found for basal metabolism. This unpredicted result sets a challenge that can decide which of the concurrent hypotheses is the correct theory. Here, we show that both scaling laws can be deduced from a single network model. Besides the 3/4-law for basal metabolism, the model predicts that maximum metabolic rate scales as M, maximum heart rate as M, and muscular capillary density as M, in agreement with data.

  7. Effects of calibration methods on quantitative material decomposition in photon-counting spectral computed tomography using a maximum a posteriori estimator.

    Science.gov (United States)

    Curtis, Tyler E; Roeder, Ryan K

    2017-10-01

    Advances in photon-counting detectors have enabled quantitative material decomposition using multi-energy or spectral computed tomography (CT). Supervised methods for material decomposition utilize an estimated attenuation for each material of interest at each photon energy level, which must be calibrated based upon calculated or measured values for known compositions. Measurements using a calibration phantom can advantageously account for system-specific noise, but the effect of calibration methods on the material basis matrix and subsequent quantitative material decomposition has not been experimentally investigated. Therefore, the objective of this study was to investigate the influence of the range and number of contrast agent concentrations within a modular calibration phantom on the accuracy of quantitative material decomposition in the image domain. Gadolinium was chosen as a model contrast agent in imaging phantoms, which also contained bone tissue and water as negative controls. The maximum gadolinium concentration (30, 60, and 90 mM) and total number of concentrations (2, 4, and 7) were independently varied to systematically investigate effects of the material basis matrix and scaling factor calibration on the quantitative (root mean squared error, RMSE) and spatial (sensitivity and specificity) accuracy of material decomposition. Images of calibration and sample phantoms were acquired using a commercially available photon-counting spectral micro-CT system with five energy bins selected to normalize photon counts and leverage the contrast agent k-edge. Material decomposition of gadolinium, calcium, and water was performed for each calibration method using a maximum a posteriori estimator. Both the quantitative and spatial accuracy of material decomposition were most improved by using an increased maximum gadolinium concentration (range) in the basis matrix calibration; the effects of using a greater number of concentrations were relatively small in

  8. Positron imaging system with improved count rate and tomographic capability

    International Nuclear Information System (INIS)

    Muehllehner, G.; Buchin, M.P.

    1980-01-01

    Improvements to a positron camera imaging system are described. A pair of Angear-type scintillation cameras serve as the detectors, each camera being positioned on opposite sides of the organ of interest. Pulse shaping circuits reduce the pulse duration below 900 nanoseconds and the integration time below 500 noneseconds, improving the count rate capability and the counting statistics of the system and thus the image quality and processing speed. The invention also provides means for rotating the opposed camera heads about an axis which passes through the organ of interest. The cameras do not use collimators, and are capable of accepting radiation travelling in planes not perpendicular to the scintillation crystals. (LL)

  9. Rate maximum calculation of Dpa in CNA-II pressure vessel

    International Nuclear Information System (INIS)

    Mascitti, J. A

    2012-01-01

    The maximum dpa rate was calculated for the reactor in the following state: fresh fuel, no Xenon, a Boron concentration of 15.3 ppm, critical state, its control rods in the criticality position, hot, at full power (2160 MW). It was determined that the maximum dpa rate under such conditions is 3.54(2)x10 12 s -1 and it is located in the positions corresponding to θ=210 o in the azimuthal direction, and z=20 cm and -60 cm respectively in the axial direction, considering the calculation mesh centered at half height of the fuel element (FE) active length. The dpa rate spectrum was determined as well as the contribution to it for 4 energy groups: a thermal group, two epithermal groups and a fast one. The maximum dpa rate considering the photo-neutrons production from (γ, n) reaction in the heavy water of coolant and moderator was 3.93(4)x10 12 s -1 that is 11% greater than the obtained without photo-neutrons. This verified significant difference between both cases, suggest that photo-neutrons in large heavy water reactors such as CNA-II should not be ignored. The maximum DPA rate in the first mm of the reactor pressure vessel was calculated too and it was obtained a value of 4.22(6)x10 12 s -1 . It should be added that the calculation was carried out with the reactor complete accurate model, with no approximations in spatial or energy variables. Each value has, between parentheses, a percentage relative error representing the statistical uncertainty due to the probabilistic Monte Carlo method used to estimate it. More representative values may be obtained with this method if equilibrium burn-up distribution is used (author)

  10. Precise method for correcting count-rate losses in scintillation cameras

    International Nuclear Information System (INIS)

    Madsen, M.T.; Nickles, R.J.

    1986-01-01

    Quantitative studies performed with scintillation detectors often require corrections for lost data because of the finite resolving time of the detector. Methods that monitor losses by means of a reference source or pulser have unacceptably large statistical fluctuations associated with their correction factors. Analytic methods that model the detector as a paralyzable system require an accurate estimate of the system resolving time. Because the apparent resolving time depends on many variables, including the window setting, source distribution, and the amount of scattering material, significant errors can be introduced by relying on a resolving time obtained from phantom measurements. These problems can be overcome by curve-fitting the data from a reference source to a paralyzable model in which the true total count rate in the selected window is estimated from the observed total rate. The resolving time becomes a free parameter in this method which is optimized to provide the best fit to the observed reference data. The fitted curve has the inherent accuracy of the reference source method with the precision associated with the observed total image count rate. Correction factors can be simply calculated from the ratio of the true reference source rate and the fitted curve. As a result, the statistical uncertainty of the data corrected by this method is not significantly increased

  11. A pulse shape discriminator with high precision of neutron and gamma ray selection at high counting rate

    International Nuclear Information System (INIS)

    Bialkowski, J.; Moszynski, M.; Wolski, D.

    1989-01-01

    A pulse shape discriminator based on the zero-crossing principle is described. Due to dc negative feedback loops stabilizing the shaping amplifier and the zero-crossing discriminator, the working of the circuit is not affected by the high counting rate and the temperature variations. The pileup rejection circuit built into the discriminator improves the quality of the n-γ separation at high counting rates. A full γ-ray rejection is obtained for a recoil energy of electrons down to 25 keV. At high counting rates the remaining γ-ray contribution is evidently due to the pileup effect which is equal to about 2% at 4x10 5 counts/s. (orig.)

  12. ULY JUP GRB SOLAR X-RAY/COSMIC GAMMA-RAY RAW COUNT RATE

    Data.gov (United States)

    National Aeronautics and Space Administration — The archived data consist of count rates from the sum of two hemispherical detectors covering 4 pi steradians and operating continuously. The detectors are 3 mm...

  13. CASA-Mot technology: how results are affected by the frame rate and counting chamber.

    Science.gov (United States)

    Bompart, Daznia; García-Molina, Almudena; Valverde, Anthony; Caldeira, Carina; Yániz, Jesús; Núñez de Murga, Manuel; Soler, Carles

    2018-04-04

    For over 30 years, CASA-Mot technology has been used for kinematic analysis of sperm motility in different mammalian species, but insufficient attention has been paid to the technical limitations of commercial computer-aided sperm analysis (CASA) systems. Counting chamber type and frame rate are two of the most important aspects to be taken into account. Counting chambers can be disposable or reusable, with different depths. In human semen analysis, reusable chambers with a depth of 10µm are the most frequently used, whereas for most farm animal species it is more common to use disposable chambers with a depth of 20µm . The frame rate was previously limited by the hardware, although changes in the number of images collected could lead to significant variations in some kinematic parameters, mainly in curvilinear velocity (VCL). A frame rate of 60 frames s-1 is widely considered to be the minimum necessary for satisfactory results. However, the frame rate is species specific and must be defined in each experimental condition. In conclusion, we show that the optimal combination of frame rate and counting chamber type and depth should be defined for each species and experimental condition in order to obtain reliable results.

  14. Positron imaging system with improved count rate and tomographic capability

    International Nuclear Information System (INIS)

    1979-01-01

    A system with improved count rate capability for detecting the radioactive distribution of positron events within an organ of interest in a living subject is described. Objects of the invention include improving the scintillation crystal and pulse processing electronics, avoiding the limitations of collimators and provide an Arger camera positron imaging system that avoids the use of collimators. (U.K.)

  15. Exercise-induced maximum metabolic rate scaled to body mass by ...

    African Journals Online (AJOL)

    user

    2016-10-27

    Oct 27, 2016 ... maximum aerobic metabolic rate (MMR) is proportional to the fractal extent ... metabolic rate with body mass can be obtained by taking body .... blood takes place. ..... MMR and BMR is that MMR is owing mainly to respiration in skeletal .... the spectra of surface area scaling strategies of cells and organisms:.

  16. CD4+ count and Nitro-Blue Tetrazolium reduction rate of neutrophil ...

    African Journals Online (AJOL)

    CD4+ count and Nitro-Blue Tetrazolium reduction rate of neutrophil in newly diagnosed HIV-infected adults in Sokoto Metropolis. U.K. Mustapha, C.C. Onyenekwe, A. Yakubu, B.R. Alkali, M.H. Yeldu, K.M. Hamid, I. Abdullahi, N.M. Bunza, M. Bello, A.B. Ibrahim ...

  17. Effect of recirculation and regional counting rate on reliability of noninvasive bicompartmental CBF measurements

    International Nuclear Information System (INIS)

    Herholz, K.

    1985-01-01

    Based on data from routine intravenous Xe133-rCBF studies in 50 patients, using Obrist's algorithm the effect of counting rate statistics and amount of recirculating activity on reproducibility of results was investigated at five simulated counting rate levels. Dependence of the standard deviation of compartmental and noncompartmental flow parameters on recirculation and counting rate was determined by multiple linear regression analysis. Those regression equations permit determination of the optimum accuracy that may be expected from individual flow measurements. Mainly due to a delay of the start-of-fit time an exponential increase in standard deviation of flow measurements was observed as recirculation increased. At constant start-of-fit, however, a linear increase in standard deviation of compartmental flow parameters only was found, while noncompartmental results remained constant. Therefore, and in regard to other studies of potential sources of error, an upper limit of 2.5 min for the start-of-fit time and usage of noncompartmental flow parameters for measurements affected by high recirculation are suggested

  18. Resolving time of scintillation camera-computer system and methods of correction for counting loss, 2

    International Nuclear Information System (INIS)

    Iinuma, Takeshi; Fukuhisa, Kenjiro; Matsumoto, Toru

    1975-01-01

    Following the previous work, counting-rate performance of camera-computer systems was investigated for two modes of data acquisition. The first was the ''LIST'' mode in which image data and timing signals were sequentially stored on magnetic disk or tape via a buffer memory. The second was the ''HISTOGRAM'' mode in which image data were stored in a core memory as digital images and then the images were transfered to magnetic disk or tape by the signal of frame timing. Firstly, the counting-rates stored in the buffer memory was measured as a function of display event-rates of the scintillation camera for the two modes. For both modes, stored counting-rated (M) were expressed by the following formula: M=N(1-Ntau) where N was the display event-rates of the camera and tau was the resolving time including analog-to-digital conversion time and memory cycle time. The resolving time for each mode may have been different, but it was about 10 μsec for both modes in our computer system (TOSBAC 3400 model 31). Secondly, the date transfer speed from the buffer memory to the external memory such as magnetic disk or tape was considered for the two modes. For the ''LIST'' mode, the maximum value of stored counting-rates from the camera was expressed in terms of size of the buffer memory, access time and data transfer-rate of the external memory. For the ''HISTOGRAM'' mode, the minimum time of the frame was determined by size of the buffer memory, access time and transfer rate of the external memory. In our system, the maximum value of stored counting-rates were about 17,000 counts/sec. with the buffer size of 2,000 words, and minimum frame time was about 130 msec. with the buffer size of 1024 words. These values agree well with the calculated ones. From the author's present analysis, design of the camera-computer system becomes possible for quantitative dynamic imaging and future improvements are suggested. (author)

  19. A compact 7-cell Si-drift detector module for high-count rate X-ray spectroscopy.

    Science.gov (United States)

    Hansen, K; Reckleben, C; Diehl, I; Klär, H

    2008-05-01

    A new Si-drift detector module for fast X-ray spectroscopy experiments was developed and realized. The Peltier-cooled module comprises a sensor with 7 × 7-mm 2 active area, an integrated circuit for amplification, shaping and detection, storage, and derandomized readout of signal pulses in parallel, and amplifiers for line driving. The compactness and hexagonal shape of the module with a wrench size of 16mm allow very short distances to the specimen and multi-module arrangements. The power dissipation is 186mW. At a shaper peaking time of 190 ns and an integration time of 450 ns an electronic rms noise of ~11 electrons was achieved. When operated at 7 °C, FWHM line widths around 260 and 460 eV (Cu-K α ) were obtained at low rates and at sum-count rates of 1.7 MHz, respectively. The peak shift is below 1% for a broad range of count rates. At 1.7-MHz sum-count rate the throughput loss amounts to 30%.

  20. A gamma camera count rate saturation correction method for whole-body planar imaging

    Science.gov (United States)

    Hobbs, Robert F.; Baechler, Sébastien; Senthamizhchelvan, Srinivasan; Prideaux, Andrew R.; Esaias, Caroline E.; Reinhardt, Melvin; Frey, Eric C.; Loeb, David M.; Sgouros, George

    2010-02-01

    Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating

  1. A count rate model for PET and its application to an LSO HR PLUS scanner

    International Nuclear Information System (INIS)

    Moisan, C.; Rogers, J.G.; Douglas, J.L.

    1996-10-01

    We present a count rate model for PET. Considering a standard 20 x 20 cm phantom in the field-of-view of a cylindrical septaless tomograph, the model computes the acceptance to prompt and random events from simple geometric considerations. Dead time factors at all stages of a typical event acquisition architecture are calculated from specified processing clock cycles. Validations of the model's predictions against the measured performances of the ECAT-953B and the EXACT HR PLUS are presented. The model is then used to investigate the benefit of using detectors made of LSO in the EXACT HR PLUS scanner geometry. The results indicate that in replacing BGO by the faster LSO, one can count on an increase of the peak noise-equivalent-count rate by a factor 2.2. This gain will be achieved by using a 5 nsec coincidence window, buckets operating on 128 nsec clock cycle, and front-end data acquisition that can sustain a total rate of 2.9 MHz. (authors)

  2. Count rate balance method of measuring sediment transport of sand beds by radioactive tracers

    International Nuclear Information System (INIS)

    Sauzay, G.

    1968-01-01

    Radioactive tracers are applied to the direct measurement of the sediment transport rate of sand beds. The theoretical measurement formula is derived: the variation of the count rate balance is inverse of that of the transport thickness. Simultaneously the representativeness of the tracer is critically studied. The minimum quantity of tracer which has to be injected in order to obtain a correct statistical definition of count rate given by a low number of grains 'seen' by the detector is then studied. A field experiment was made and has let to study the technological conditions for applying this method: only the treatment of results is new, the experiment itself is carried out with conventional techniques applied with great care. (author) [fr

  3. Estimation of maximum credible atmospheric radioactivity concentrations and dose rates from nuclear tests

    International Nuclear Information System (INIS)

    Telegadas, K.

    1979-01-01

    A simple technique is presented for estimating maximum credible gross beta air concentrations from nuclear detonations in the atmosphere, based on aircraft sampling of radioactivity following each Chinese nuclear test from 1964 to 1976. The calculated concentration is a function of the total yield and fission yield, initial vertical radioactivity distribution, time after detonation, and rate of horizontal spread of the debris with time. calculated maximum credible concentrations are compared with the highest concentrations measured during aircraft sampling. The technique provides a reasonable estimate of maximum air concentrations from 1 to 10 days after a detonation. An estimate of the whole-body external gamma dose rate corresponding to the maximum credible gross beta concentration is also given. (author)

  4. Optimizing the calculation of point source count-centroid in pixel size measurement

    International Nuclear Information System (INIS)

    Zhou Luyi; Kuang Anren; Su Xianyu

    2004-01-01

    Purpose: Pixel size is an important parameter of gamma camera and SPECT. A number of Methods are used for its accurate measurement. In the original count-centroid method, where the image of a point source(PS) is acquired and its count-centroid calculated to represent PS position in the image, background counts are inevitable. Thus the measured count-centroid (Xm) is an approximation of the true count-centroid (Xp) of the PS, i.e. Xm=Xp+(Xb-Xp)/(1+Rp/Rb), where Rp is the net counting rate of the PS, Xb the background count-centroid and Rb the background counting rate. To get accurate measurement, Rp must be very big, which is unpractical, resulting in the variation of measured pixel size. Rp-independent calculation of PS count-centroid is desired. Methods: The proposed method attempted to eliminate the effect of the term (Xb-Xp)/(1+Rp/Rb) by bringing Xb closer to Xp and by reducing Rb. In the acquired PS image, a circular ROI was generated to enclose the PS, the pixel with the maximum count being the center of the ROI. To choose the diameter (D) of the ROI, a Gaussian count distribution was assumed for the PS, accordingly, K=I-(0.5)D/R percent of the total PS counts was in the ROI, R being the full width at half maximum of the PS count distribution. D was set to be 6*R to enclose most (K=98.4%) of the PS counts. The count-centroid of the ROI was calculated to represent Xp. The proposed method was tested in measuring the pixel size of a well-tuned SPECT, whose pixel size was estimated to be 3.02 mm according to its mechanical and electronic setting (128*128 matrix, 387 mm UFOV, ZOOM=1). For comparison, the original method, which was use in the former versions of some commercial SPECT software, was also tested. 12 PSs were prepared and their image acquired and stored. The net counting rate of the PSs increased from 10cps to 1183cps. Results: Using the proposed method, the measured pixel size (in mm) varied only between 3.00 and 3.01( mean= 3.01±0.00) as Rp increased

  5. On the counting rate-dependent amplitude shifts in proportional counters

    International Nuclear Information System (INIS)

    Mahesh, K.

    1976-01-01

    The presence of a positive ion cloud near the anode wire, which has been believed to be the cause of the observed counting rate-dependent peak shifts in proportional counters, does not explain various exhibited features of these shifts. The possibility of other mechanisms involved is, therefore, examined. The columnar recombination of primary ions in the counter is considered to explain the origin of the shifts and their observed features. (Auth.)

  6. Development of Fast High-Resolution Muon Drift-Tube Detectors for High Counting Rates

    CERN Document Server

    INSPIRE-00287945; Dubbert, J.; Horvat, S.; Kortner, O.; Kroha, H.; Legger, F.; Richter, R.; Adomeit, S.; Biebel, O.; Engl, A.; Hertenberger, R.; Rauscher, F.; Zibell, A.

    2011-01-01

    Pressurized drift-tube chambers are e?cient detectors for high-precision tracking over large areas. The Monitored Drift-Tube (MDT) chambers of the muon spectrometer of the ATLAS detector at the Large Hadron Collider (LHC) reach a spatial resolution of 35 micons and almost 100% tracking e?ciency with 6 layers of 30 mm diameter drift tubes operated with Ar:CO2 (93:7) gas mixture at 3 bar and a gas gain of 20000. The ATLAS MDT chambers are designed to cope with background counting rates due to neutrons and gamma-rays of up to about 300 kHz per tube which will be exceeded for LHC luminosities larger than the design value of 10-34 per square cm and second. Decreasing the drift-tube diameter to 15 mm while keeping the other parameters, including the gas gain, unchanged reduces the maximum drift time from about 700 ns to 200 ns and the drift-tube occupancy by a factor of 7. New drift-tube chambers for the endcap regions of the ATLAS muon spectrometer have been designed. A prototype chamber consisting of 12 times 8 l...

  7. Maximum production rate optimization for sulphuric acid decomposition process in tubular plug-flow reactor

    International Nuclear Information System (INIS)

    Wang, Chao; Chen, Lingen; Xia, Shaojun; Sun, Fengrui

    2016-01-01

    A sulphuric acid decomposition process in a tubular plug-flow reactor with fixed inlet flow rate and completely controllable exterior wall temperature profile and reactants pressure profile is studied in this paper by using finite-time thermodynamics. The maximum production rate of the aimed product SO 2 and the optimal exterior wall temperature profile and reactants pressure profile are obtained by using nonlinear programming method. Then the optimal reactor with the maximum production rate is compared with the reference reactor with linear exterior wall temperature profile and the optimal reactor with minimum entropy generation rate. The result shows that the production rate of SO 2 of optimal reactor with the maximum production rate has an increase of more than 7%. The optimization of temperature profile has little influence on the production rate while the optimization of reactants pressure profile can significantly increase the production rate. The results obtained may provide some guidelines for the design of real tubular reactors. - Highlights: • Sulphuric acid decomposition process in tubular plug-flow reactor is studied. • Fixed inlet flow rate and controllable temperature and pressure profiles are set. • Maximum production rate of aimed product SO 2 is obtained. • Corresponding optimal temperature and pressure profiles are derived. • Production rate of SO 2 of optimal reactor increases by 7%.

  8. High resolution gamma-ray spectroscopy at high count rates with a prototype High Purity Germanium detector

    Science.gov (United States)

    Cooper, R. J.; Amman, M.; Vetter, K.

    2018-04-01

    High-resolution gamma-ray spectrometers are required for applications in nuclear safeguards, emergency response, and fundamental nuclear physics. To overcome one of the shortcomings of conventional High Purity Germanium (HPGe) detectors, we have developed a prototype device capable of achieving high event throughput and high energy resolution at very high count rates. This device, the design of which we have previously reported on, features a planar HPGe crystal with a reduced-capacitance strip electrode geometry. This design is intended to provide good energy resolution at the short shaping or digital filter times that are required for high rate operation and which are enabled by the fast charge collection afforded by the planar geometry crystal. In this work, we report on the initial performance of the system at count rates up to and including two million counts per second.

  9. Maximum discharge rate of liquid-vapor mixtures from vessels

    International Nuclear Information System (INIS)

    Moody, F.J.

    1975-09-01

    A discrepancy exists in theoretical predictions of the two-phase equilibrium discharge rate from pipes attached to vessels. Theory which predicts critical flow data in terms of pipe exit pressure and quality severely overpredicts flow rates in terms of vessel fluid properties. This study shows that the discrepancy is explained by the flow pattern. Due to decompression and flashing as fluid accelerates into the pipe entrance, the maximum discharge rate from a vessel is limited by choking of a homogeneous bubbly mixture. The mixture tends toward a slip flow pattern as it travels through the pipe, finally reaching a different choked condition at the pipe exit

  10. Pulse-duration discrimination for increasing counting characteristic plateau and for improving counting rate stability of a scintillation counter

    International Nuclear Information System (INIS)

    Kuz'min, M.G.

    1977-01-01

    For greater stability of scintillation counters operation, discussed is the possibility for increasing the plateau and reducing its slope. Presented is the circuit for discrimination of the signal pulses from input pulses of a photomultiplier. The counting characteristics have been measured with the scintillation detectors being irradiated by different gamma sources ( 60 Co, 137 Cs, 241 Am) and without the source when the scintillation detector is shielded by a tungsten cylinder with a wall thickness of 23 mm. The comparison has revealed that discrimination in duration increase the plateau and reduces its slope. Proceeding from comparison of the noise characteristics, the relationship is found between the noise pulse number and gamma radiation energy. For better stability of the counting rate it is suggested to introduce into the scintillation counter the circuit for duration discrimination of the output pulses of a photomultiplier

  11. Ultra-fast photon counting with a passive quenching silicon photomultiplier in the charge integration regime

    Science.gov (United States)

    Zhang, Guoqing; Lina, Liu

    2018-02-01

    An ultra-fast photon counting method is proposed based on the charge integration of output electrical pulses of passive quenching silicon photomultipliers (SiPMs). The results of the numerical analysis with actual parameters of SiPMs show that the maximum photon counting rate of a state-of-art passive quenching SiPM can reach ~THz levels which is much larger than that of the existing photon counting devices. The experimental procedure is proposed based on this method. This photon counting regime of SiPMs is promising in many fields such as large dynamic light power detection.

  12. On the equivalence between the minimum entropy generation rate and the maximum conversion rate for a reactive system

    International Nuclear Information System (INIS)

    Bispo, Heleno; Silva, Nilton; Brito, Romildo; Manzi, João

    2013-01-01

    Highlights: • Minimum entropy generation (MEG) principle improved the reaction performance. • MEG rate and the maximum conversion equivalence have been analyzed. • Temperature and residence time are used to the domain establishment of MEG. • Satisfying the temperature and residence time relationship results a optimal performance. - Abstract: The analysis of the equivalence between the minimum entropy generation (MEG) rate and the maximum conversion rate for a reactive system is the main purpose of this paper. While being used as a strategy of optimization, the minimum entropy production was applied to the production of propylene glycol in a Continuous Stirred-Tank Reactor (CSTR) with a view to determining the best operating conditions, and under such conditions, a high conversion rate was found. The effects of the key variables and restrictions on the validity domain of MEG were investigated, which raises issues that are included within a broad discussion. The results from simulations indicate that from the chemical reaction standpoint a maximum conversion rate can be considered as equivalent to MEG. Such a result can be clearly explained by examining the classical Maxwell–Boltzmann distribution, where the molecules of the reactive system under the condition of the MEG rate present a distribution of energy with reduced dispersion resulting in a better quality of collision between molecules with a higher conversion rate

  13. An integral whole circuit of amplifying and discriminating suited to high counting rate

    International Nuclear Information System (INIS)

    Dong Chengfu; Su Hong; Wu Ming; Li Xiaogang; Peng Yu; Qian Yi; Liu Yicai; Xu Sijiu; Ma Xiaoli

    2007-01-01

    A hybrid circuit consists of charge sensitive preamplifier, main amplifier, discriminator and shaping circuit was described. This instrument has characteristics of low power consumption, small volume, high sensitivity, potable and so on, and is convenient for use in field. The output pulse of this instrument may directly consist with CMOS or TTL logic level. This instrument was mainly used for count measurement, for example, for high sensitive 3 He neutron detector, meanwhile also may used for other heavy ion detectors, the highest counting rate can reach 10 6 /s. (authors)

  14. High counting rate resistive-plate chamber

    International Nuclear Information System (INIS)

    Peskov, V.; Anderson, D.F.; Kwan, S.

    1993-05-01

    Parallel-plate avalanche chambers (PPAC) are widely used in physics experiments because they are fast ( 5 counts/mm 2 . A resistive-plate chamber (RPC) is similar to the PPAC in construction except that one or both of the electrodes are made from high resistivity (≥10 10 Ω·cm) materials. In practice RPCs are usually used in the spark mode. Resistive electrodes are charged by sparks, locally reducing the actual electric field in the gap. The size of the charged surface is about 10 mm 2 , leaving the rest of the detector unaffected. Therefore, the rate capability of such detectors in the spark mode is considerably higher than conventional spark counters. Among the different glasses tested the best results were obtained with electron type conductive glasses, which obey Ohm's law. Most of the work with such glasses was done with high pressure parallel-plate chambers (10 atm) for time-of-flight measurements. Resistive glasses have been expensive and produced only in small quantities. Now resistive glasses are commercially available, although they are still expensive in small scale production. From the positive experience of different groups working with the resistive glasses, it was decided to review the old idea to use this glass for the RPC. This work has investigated the possibility of using the RPC at 1 atm and in the avalanche mode. This has several advantages: simplicity of construction, high rate capability, low voltage operation, and the ability to work with non-flammable gases

  15. Fly's Eye: a counting camera for thermal neutrons, some applications, problems, and prospects

    International Nuclear Information System (INIS)

    Davidson, J.B.

    1975-01-01

    An area detector for thermal neutrons based on image intensification techniques is described and some capabilities and limitations of the detection system are discussed. Among the former are high spatial resolution high instantaneous counting rate, electronic zoom, time-gating, and integration. The detector is limited in that the maximum counting rate for a resolution element is 60 regularly spaced counts per second. Also, the nonuniformity of response over the detector limits the useful size and requires point-by-point calibration. In addition, a higher efficiency for neutron detection would be desirable. Some typical applications of the system are: crystal inspection, neutron magnetic diffraction topography, and searches for temperature induced changes in diffraction patterns. The future application of solid state television sensors and microchannel plate intensifiers to improve the system are briefly mentioned. (U.S.)

  16. Fly's eye: a counting camera for thermal neutrons: some applications, problems, and prospects

    International Nuclear Information System (INIS)

    Davidson, J.B.

    1976-01-01

    An area detector for thermal neutrons based on image intensification techniques is described. Some capabilities and limitations of the detection system are discussed. Among the former are high spatial resolution, high instantaneous counting rate, electronic zoom, time-gating, and integration. The detector is limited in that the maximum counting rate for a resolution element is 60 regularly spaced counts per second. Also, the nonuniformity of response over the detector puts a limit on the useful size and necessitates point-by-point calibration. In addition, a higher efficiency for neutron detection would be desirable. Some typical applications of the system are crystal inspection, neutron magnetic diffraction topography, and searches for temperature-induced changes in diffraction patterns. The future application of solid-state television sensors and microchannel-plate intensifiers to improve the system is briefly mentioned

  17. Analysis of geomagnetic storm variations and count-rate of cosmic ray muons recorded at the Brazilian southern space observatory

    International Nuclear Information System (INIS)

    Frigo, Everton; Savian, Jairo Francisco; Silva, Marlos Rockenbach da; Lago, Alisson dal; Trivedi, Nalin Babulal; Schuch, Nelson Jorge

    2007-01-01

    An analysis of geomagnetic storm variations and the count rate of cosmic ray muons recorded at the Brazilian Southern Space Observatory -OES/CRS/INPE-MCT, in Sao Martinho da Serra, RS during the month of November 2004, is presented in this paper. The geomagnetic measurements are done by a three component low noise fluxgate magnetometer and the count rates of cosmic ray muons are recorded by a muon scintillator telescope - MST, both instruments installed at the Observatory. The fluxgate magnetometer measures variations in the three orthogonal components of Earth magnetic field, H (North-South), D (East-West) and Z (Vertical), with data sampling rate of 0.5 Hz. The muon scintillator telescope records hourly count rates. The arrival of a solar disturbance can be identified by observing the decrease in the muon count rate. The goal of this work is to describe the physical morphology and phenomenology observed during the geomagnetic storm of November 2004, using the H component of the geomagnetic field and vertical channel V of the multi-directional muon detector in South of Brazil. (author)

  18. Analysis of geomagnetic storm variations and count-rate of cosmic ray muons recorded at the Brazilian southern space observatory

    Energy Technology Data Exchange (ETDEWEB)

    Frigo, Everton [University of Sao Paulo, USP, Institute of Astronomy, Geophysics and Atmospheric Sciences, IAG/USP, Department of Geophysics, Sao Paulo, SP (Brazil); Savian, Jairo Francisco [Space Science Laboratory of Santa Maria, LACESM/CT, Southern Regional Space Research Center, CRS/INPE, MCT, Santa Maria, RS (Brazil); Silva, Marlos Rockenbach da; Lago, Alisson dal; Trivedi, Nalin Babulal [National Institute for Space Research, INPE/MCT, Division of Space Geophysics, DGE, Sao Jose dos Campos, SP (Brazil); Schuch, Nelson Jorge, E-mail: efrigo@iag.usp.br, E-mail: savian@lacesm.ufsm.br, E-mail: njschuch@lacesm.ufsm.br, E-mail: marlos@dge.inpe.br, E-mail: dallago@dge.inpe.br, E-mail: trivedi@dge.inpe.br [Southern Regional Space Research Center, CRS/INPE, MCT, Santa Maria, RS (Brazil)

    2007-07-01

    An analysis of geomagnetic storm variations and the count rate of cosmic ray muons recorded at the Brazilian Southern Space Observatory -OES/CRS/INPE-MCT, in Sao Martinho da Serra, RS during the month of November 2004, is presented in this paper. The geomagnetic measurements are done by a three component low noise fluxgate magnetometer and the count rates of cosmic ray muons are recorded by a muon scintillator telescope - MST, both instruments installed at the Observatory. The fluxgate magnetometer measures variations in the three orthogonal components of Earth magnetic field, H (North-South), D (East-West) and Z (Vertical), with data sampling rate of 0.5 Hz. The muon scintillator telescope records hourly count rates. The arrival of a solar disturbance can be identified by observing the decrease in the muon count rate. The goal of this work is to describe the physical morphology and phenomenology observed during the geomagnetic storm of November 2004, using the H component of the geomagnetic field and vertical channel V of the multi-directional muon detector in South of Brazil. (author)

  19. Fumigant dosages below maximum label rate control some soilborne pathogens

    Directory of Open Access Journals (Sweden)

    Shachaf Triky-Dotan

    2016-08-01

    Full Text Available The activity of commercial soil fumigants on some key soilborne pathogens was assessed in sandy loam soil under controlled conditions. Seven soil fumigants that are registered in California or are being or have been considered for registration were used in this study: dimethyl disulfide (DMDS mixed with chloropicrin (Pic (79% DMDS and 21% Pic, Tri-Con (50% methyl bromide and 50% Pic, Midas Gold (33% methyl iodide [MI] and 67% Pic, Midas Bronze (50% MI and 50% Pic, Midas (MI, active ingredient [a.i.] 97.8%, Pic (a.i. 99% trichloronitromethane and Pic-Clor 60 (57% Pic and 37% 1,3-dichloropropene [1–3,D]. Dose-response models were calculated for pathogen mortality after 24 hours of exposure to fumigants. Overall, the tested fumigants achieved good efficacy with dosages below the maximum label rate against the tested pathogens. In this study, Pythium ultimum and citrus nematode were sensitive to all the fumigants and Verticillium dahliae was resistant. For most fumigants, California regulations restrict application rates to less than the maximum (federal label rate, meaning that it is possible that the fumigants may not control major plant pathogens. This research provides information on the effectiveness of these alternatives at these lower application rates. The results from this study will help growers optimize application rates for registered fumigants (such as Pic and 1,3-D and will help accelerate the adoption of new fumigants (such as DMDS if they are registered in California.

  20. Expected count rate for the Self- Interrogation Neutron Resonance Densitometry measurements of spent nuclear fuel

    Energy Technology Data Exchange (ETDEWEB)

    Rossa, Riccardo [SCK.CEN, Belgian Nuclear Research Centre, Boeretang, 200 - B2400 Mol (Belgium); Universite libre de Bruxelles, Ecole polytechnique de Bruxelles - Service de Metrologie Nucleaire, CP 165/84, Avenue F.D. Roosevelt, 50 - B1050 Brussels (Belgium); Borella, Alessandro; Van der Meer, Klaas [SCK.CEN, Belgian Nuclear Research Centre, Boeretang, 200 - B2400 Mol (Belgium); Labeau, Pierre-Etienne; Pauly, Nicolas [Universite libre de Bruxelles, Ecole polytechnique de Bruxelles - Service de Metrologie Nucleaire, CP 165/84, Avenue F.D. Roosevelt, 50 - B1050 Brussels (Belgium)

    2015-07-01

    The Self-Interrogation Neutron Resonance Densitometry (SINRD) is a passive neutron technique that aims at a direct quantification of {sup 239}Pu in the fuel assemblies by measuring the attenuation of the neutron flux in the energy region close to the 0.3 eV resonance of {sup 239}Pu. The {sup 239}Pu mass is estimated by calculating the SINRD signature, that is the ratio between the neutron flux integrated over the fast energy region and around the 0.3 eV resonance region. The SINRD measurement approach considered in this study consists in introducing a small neutron detector in the central guide tube of a PWR 17x17 fuel assembly. In order to measure the neutron flux in the energy regions defined in the SINRD signature, different detector types are used. The response of a bare {sup 238}U fission chamber is considered for the determination of the fast neutron flux, while other thermal-epithermal detectors wrapped in neutron absorbers are envisaged to measure the neutron flux around the resonance region. This paper provides an estimation of the count rate that can be achieved with the detector types proposed for the SINRD measurement. In the first section a set of detectors are evaluated in terms of count rate and sensitivity to the {sup 239}Pu content, in order to identify the optimal measurement configuration for each detector type. Then a study is performed to increase the count rate by increasing the detector size. The study shows that the highest count rate is achieved by using either {sup 3}He or {sup 10}B proportional counters because of the high neutron efficiency of these detectors. However, the calculations indicate that the biggest contribution to the measurement uncertainty is due to the measurement of the fast neutron flux. Finally, similar sensitivity to the {sup 239}Pu content is obtained by using the different detector types for the measurement of the neutron flux close to the resonance region. Therefore, the count rate associated to each detector type

  1. [The maximum heart rate in the exercise test: the 220-age formula or Sheffield's table?].

    Science.gov (United States)

    Mesquita, A; Trabulo, M; Mendes, M; Viana, J F; Seabra-Gomes, R

    1996-02-01

    To determine in the maximum cardiac rate in exercise test of apparently healthy individuals may be more properly estimated through 220-age formula (Astrand) or the Sheffield table. Retrospective analysis of clinical history and exercises test of apparently healthy individuals submitted to cardiac check-up. Sequential sampling of 170 healthy individuals submitted to cardiac check-up between April 1988 and September 1992. Comparison of maximum cardiac rate of individuals studied by the protocols of Bruce and modified Bruce, in interrupted exercise test by fatigue, and with the estimated values by the formulae: 220-age versus Sheffield table. The maximum cardiac heart rate is similar with both protocols. This parameter in normal individuals is better predicted by the 220-age formula. The theoretic maximum cardiac heart rate determined by 220-age formula should be recommended for a healthy, and for this reason the Sheffield table has been excluded from our clinical practice.

  2. 13 CFR 107.845 - Maximum rate of amortization on Loans and Debt Securities.

    Science.gov (United States)

    2010-01-01

    ... ADMINISTRATION SMALL BUSINESS INVESTMENT COMPANIES Financing of Small Businesses by Licensees Structuring Licensee's Financing of An Eligible Small Business: Terms and Conditions of Financing § 107.845 Maximum... 13 Business Credit and Assistance 1 2010-01-01 2010-01-01 false Maximum rate of amortization on...

  3. 2D dark-count-rate modeling of PureB single-photon avalanche diodes in a TCAD environment

    Science.gov (United States)

    Knežević, Tihomir; Nanver, Lis K.; Suligoj, Tomislav

    2018-02-01

    PureB silicon photodiodes have nm-shallow p+n junctions with which photons/electrons with penetration-depths of a few nanometer can be detected. PureB Single-Photon Avalanche Diodes (SPADs) were fabricated and analysed by 2D numerical modeling as an extension to TCAD software. The very shallow p+ -anode has high perimeter curvature that enhances the electric field. In SPADs, noise is quantified by the dark count rate (DCR) that is a measure for the number of false counts triggered by unwanted processes in the non-illuminated device. Just like for desired events, the probability a dark count increases with increasing electric field and the perimeter conditions are critical. In this work, the DCR was studied by two 2D methods of analysis: the "quasi-2D" (Q-2D) method where vertical 1D cross-sections were assumed for calculating the electron/hole avalanche-probabilities, and the "ionization-integral 2D" (II-2D) method where crosssections were placed where the maximum ionization-integrals were calculated. The Q-2D method gave satisfactory results in structures where the peripheral regions had a small contribution to the DCR, such as in devices with conventional deepjunction guard rings (GRs). Otherwise, the II-2D method proved to be much more precise. The results show that the DCR simulation methods are useful for optimizing the compromise between fill-factor and p-/n-doping profile design in SPAD devices. For the experimentally investigated PureB SPADs, excellent agreement of the measured and simulated DCR was achieved. This shows that although an implicit GR is attractively compact, the very shallow pn-junction gives a risk of having such a low breakdown voltage at the perimeter that the DCR of the device may be negatively impacted.

  4. Maximum Likelihood Time-of-Arrival Estimation of Optical Pulses via Photon-Counting Photodetectors

    Science.gov (United States)

    Erkmen, Baris I.; Moision, Bruce E.

    2010-01-01

    Many optical imaging, ranging, and communications systems rely on the estimation of the arrival time of an optical pulse. Recently, such systems have been increasingly employing photon-counting photodetector technology, which changes the statistics of the observed photocurrent. This requires time-of-arrival estimators to be developed and their performances characterized. The statistics of the output of an ideal photodetector, which are well modeled as a Poisson point process, were considered. An analytical model was developed for the mean-square error of the maximum likelihood (ML) estimator, demonstrating two phenomena that cause deviations from the minimum achievable error at low signal power. An approximation was derived to the threshold at which the ML estimator essentially fails to provide better than a random guess of the pulse arrival time. Comparing the analytic model performance predictions to those obtained via simulations, it was verified that the model accurately predicts the ML performance over all regimes considered. There is little prior art that attempts to understand the fundamental limitations to time-of-arrival estimation from Poisson statistics. This work establishes both a simple mathematical description of the error behavior, and the associated physical processes that yield this behavior. Previous work on mean-square error characterization for ML estimators has predominantly focused on additive Gaussian noise. This work demonstrates that the discrete nature of the Poisson noise process leads to a distinctly different error behavior.

  5. Fluorescence decay data analysis correcting for detector pulse pile-up at very high count rates

    Science.gov (United States)

    Patting, Matthias; Reisch, Paja; Sackrow, Marcus; Dowler, Rhys; Koenig, Marcelle; Wahl, Michael

    2018-03-01

    Using time-correlated single photon counting for the purpose of fluorescence lifetime measurements is usually limited in speed due to pile-up. With modern instrumentation, this limitation can be lifted significantly, but some artifacts due to frequent merging of closely spaced detector pulses (detector pulse pile-up) remain an issue to be addressed. We propose a data analysis method correcting for this type of artifact and the resulting systematic errors. It physically models the photon losses due to detector pulse pile-up and incorporates the loss in the decay fit model employed to obtain fluorescence lifetimes and relative amplitudes of the decay components. Comparison of results with and without this correction shows a significant reduction of systematic errors at count rates approaching the excitation rate. This allows quantitatively accurate fluorescence lifetime imaging at very high frame rates.

  6. Total lymphocyte count as a substitute to cd4 count in management of hiv infected individuals in resource limited society

    International Nuclear Information System (INIS)

    Daud, M.Y.; Qazi, R.A.

    2015-01-01

    Pakistan is a resource limited society and gold standard parameters to monitor HIV disease activity are very costly. The objective of the study was to evaluate total lymphocyte count (TLC) as a surrogate to CD4 count to monitor disease activity in HIV/AIDS in resource limited society. Methods: This cross sectional study was carried out at HIV/AIDS treatment centre, Pakistan Institute of Medical Sciences (PIMS), Islamabad. A total of seven hundred and seventy four (774) HIV positive patients were enrolled in this study, and their CD4 count and total lymphocyte count were checked to find any correlation between the two by using Spearman ranked correlation coefficient. Results: The mean CD4 count was (434.30 ± 269.23), with minimum CD4 count of (9.00), and maximum of (1974.00). The mean total lymphocyte count (TLC) was (6764.0052 ± 2364.02) with minimum TLC (1200.00) and maximum TLC was (20200.00). Using the Pearson's correlation (r) there was a significant and positive correlation between TLC and CD4 count. (r2=0.127 and p=0.000) at 0.01 level. Conclusion: Our study showed a significant positive correlation between CD4 count and total lymphocyte count (TLC), so TLC can be used as a marker of disease activity in HIV infected patients. (author)

  7. A constant velocity Moessbauer spectrometer free of long-term instrumental drifts in the count rate

    International Nuclear Information System (INIS)

    Sarma, P.R.; Sharma, A.K.; Tripathi, K.C.

    1979-01-01

    Two new control circuits to be used with a constant velocity Moessbauer spectrometer with a loud-speaker drive have been described. The wave-forms generated in the circuits are of the stair-case type instead of the usual square wave-form, so that in each oscillation of the source it remains stationary for a fraction of the time-period. The gamma-rays counted during this period are monitored along with the positive and negative velocity counts and are used to correct any fluctuation in the count rate by feeding these pulses into the timer. The associated logic circuits have been described and the statistical errors involved in the circuits have been computed. (auth.)

  8. Methodological aspects of crossover and maximum fat-oxidation rate point determination.

    Science.gov (United States)

    Michallet, A-S; Tonini, J; Regnier, J; Guinot, M; Favre-Juvin, A; Bricout, V; Halimi, S; Wuyam, B; Flore, P

    2008-11-01

    Indirect calorimetry during exercise provides two metabolic indices of substrate oxidation balance: the crossover point (COP) and maximum fat oxidation rate (LIPOXmax). We aimed to study the effects of the analytical device, protocol type and ventilatory response on variability of these indices, and the relationship with lactate and ventilation thresholds. After maximum exercise testing, 14 relatively fit subjects (aged 32+/-10 years; nine men, five women) performed three submaximum graded tests: one was based on a theoretical maximum power (tMAP) reference; and two were based on the true maximum aerobic power (MAP). Gas exchange was measured concomitantly using a Douglas bag (D) and an ergospirometer (E). All metabolic indices were interpretable only when obtained by the D reference method and MAP protocol. Bland and Altman analysis showed overestimation of both indices with E versus D. Despite no mean differences between COP and LIPOXmax whether tMAP or MAP was used, the individual data clearly showed disagreement between the two protocols. Ventilation explained 10-16% of the metabolic index variations. COP was correlated with ventilation (r=0.96, P<0.01) and the rate of increase in blood lactate (r=0.79, P<0.01), and LIPOXmax correlated with the ventilation threshold (r=0.95, P<0.01). This study shows that, in fit healthy subjects, the analytical device, reference used to build the protocol and ventilation responses affect metabolic indices. In this population, and particularly to obtain interpretable metabolic indices, we recommend a protocol based on the true MAP or one adapted to include the transition from fat to carbohydrate. The correlation between metabolic indices and lactate/ventilation thresholds suggests that shorter, classical maximum progressive exercise testing may be an alternative means of estimating these indices in relatively fit subjects. However, this needs to be confirmed in patients who have metabolic defects.

  9. Disentangling the effects of alternation rate and maximum run length on judgments of randomness

    Directory of Open Access Journals (Sweden)

    Sabine G. Scholl

    2011-08-01

    Full Text Available Binary sequences are characterized by various features. Two of these characteristics---alternation rate and run length---have repeatedly been shown to influence judgments of randomness. The two characteristics, however, have usually been investigated separately, without controlling for the other feature. Because the two features are correlated but not identical, it seems critical to analyze their unique impact, as well as their interaction, so as to understand more clearly what influences judgments of randomness. To this end, two experiments on the perception of binary sequences orthogonally manipulated alternation rate and maximum run length (i.e., length of the longest run within the sequence. Results show that alternation rate consistently exerts a unique effect on judgments of randomness, but that the effect of alternation rate is contingent on the length of the longest run within the sequence. The effect of maximum run length was found to be small and less consistent. Together, these findings extend prior randomness research by integrating literature from the realms of perception, categorization, and prediction, as well as by showing the unique and joint effects of alternation rate and maximum run length on judgments of randomness.

  10. 19 CFR 212.07 - Rulemaking on maximum rates for attorney fees.

    Science.gov (United States)

    2010-04-01

    ... 19 Customs Duties 3 2010-04-01 2010-04-01 false Rulemaking on maximum rates for attorney fees. 212.07 Section 212.07 Customs Duties UNITED STATES INTERNATIONAL TRADE COMMISSION INVESTIGATIONS OF UNFAIR PRACTICES IN IMPORT TRADE IMPLEMENTATION OF THE EQUAL ACCESS TO JUSTICE ACT General Provisions...

  11. Optimizing the calculation of point source count-centroid in pixel size measurement

    International Nuclear Information System (INIS)

    Zhou Luyi; Kuang Anren; Su Xianyu

    2004-01-01

    Pixel size is an important parameter of gamma camera and SPECT. A number of methods are used for its accurate measurement. In the original count-centroid method, where the image of a point source (PS) is acquired and its count-centroid calculated to represent PS position in the image, background counts are inevitable. Thus the measured count-centroid (X m ) is an approximation of the true count-centroid (X p ) of the PS, i.e. X m =X p + (X b -X p )/(1+R p /R b ), where Rp is the net counting rate of the PS, X b the background count-centroid and Rb the background counting. To get accurate measurement, R p must be very big, which is unpractical, resulting in the variation of measured pixel size. R p -independent calculation of PS count-centroid is desired. Methods: The proposed method attempted to eliminate the effect of the term (X b -X p )/(1 + R p /R b ) by bringing X b closer to X p and by reducing R b . In the acquired PS image, a circular ROI was generated to enclose the PS, the pixel with the maximum count being the center of the ROI. To choose the diameter (D) of the ROI, a Gaussian count distribution was assumed for the PS, accordingly, K=1-(0.5) D/R percent of the total PS counts was in the ROI, R being the full width at half maximum of the PS count distribution. D was set to be 6*R to enclose most (K=98.4%) of the PS counts. The count-centroid of the ROI was calculated to represent X p . The proposed method was tested in measuring the pixel size of a well-tuned SPECT, whose pixel size was estimated to be 3.02 mm according to its mechanical and electronic setting (128 x 128 matrix, 387 mm UFOV, ZOOM=1). For comparison, the original method, which was use in the former versions of some commercial SPECT software, was also tested. 12 PSs were prepared and their image acquired and stored. The net counting rate of the PSs increased from 10 cps to 1183 cps. Results: Using the proposed method, the measured pixel size (in mm) varied only between 3.00 and 3.01 (mean

  12. Low power ion spectrometer for high counting rates

    International Nuclear Information System (INIS)

    Klein, J.W.; Dullenkopf, P.; Glasmachers, A.; Melbert, J.; Winkelnkemper, W.

    1980-01-01

    This report describes in detail the electronic concept for a time-of-flight (TOF) ion spectrometer for high counting rates and high dynamic range which can be used as a satellite instrument. The detection principle of the spectrometer is based on a time-of-flight and energy measurement for each incident ion. The ionmass is related to these two quantities by a simple equation. The described approach for the mass identification systems is using an analog fast-slow concept: The fast TOF-signal preselects the gainstep in the much slower energy channel. The conversion time of the mass identifier is approximately 10 -6 s and the dynamic range of the energy channel is better than 10 3 (20 keV to 25 MeV). The purpose of this study was to demonstrate the feasibility of a TOF-spectrometer capable to measure the ion composition in planetary magnetospheres. (orig.) [de

  13. Low reproducibility of maximum urinary flow rate determined by portable flowmetry

    NARCIS (Netherlands)

    Sonke, G. S.; Kiemeney, L. A.; Verbeek, A. L.; Kortmann, B. B.; Debruyne, F. M.; de la Rosette, J. J.

    1999-01-01

    To evaluate the reproducibility in maximum urinary flow rate (Qmax) in men with lower urinary tract symptoms (LUTSs) and to determine the number of flows needed to obtain a specified reliability in mean Qmax, 212 patients with LUTSs (mean age, 62 years) referred to the University Hospital Nijmegen,

  14. Characteristic performance evaluation of a photon counting Si strip detector for low dose spectral breast CT imaging

    Science.gov (United States)

    Cho, Hyo-Min; Barber, William C.; Ding, Huanjun; Iwanczyk, Jan S.; Molloi, Sabee

    2014-01-01

    Purpose: The possible clinical applications which can be performed using a newly developed detector depend on the detector's characteristic performance in a number of metrics including the dynamic range, resolution, uniformity, and stability. The authors have evaluated a prototype energy resolved fast photon counting x-ray detector based on a silicon (Si) strip sensor used in an edge-on geometry with an application specific integrated circuit to record the number of x-rays and their energies at high flux and fast frame rates. The investigated detector was integrated with a dedicated breast spectral computed tomography (CT) system to make use of the detector's high spatial and energy resolution and low noise performance under conditions suitable for clinical breast imaging. The aim of this article is to investigate the intrinsic characteristics of the detector, in terms of maximum output count rate, spatial and energy resolution, and noise performance of the imaging system. Methods: The maximum output count rate was obtained with a 50 W x-ray tube with a maximum continuous output of 50 kVp at 1.0 mA. A109Cd source, with a characteristic x-ray peak at 22 keV from Ag, was used to measure the energy resolution of the detector. The axial plane modulation transfer function (MTF) was measured using a 67 μm diameter tungsten wire. The two-dimensional (2D) noise power spectrum (NPS) was measured using flat field images and noise equivalent quanta (NEQ) were calculated using the MTF and NPS results. The image quality parameters were studied as a function of various radiation doses and reconstruction filters. The one-dimensional (1D) NPS was used to investigate the effect of electronic noise elimination by varying the minimum energy threshold. Results: A maximum output count rate of 100 million counts per second per square millimeter (cps/mm2) has been obtained (1 million cps per 100 × 100 μm pixel). The electrical noise floor was less than 4 keV. The energy resolution

  15. Improvements in the energy resolution and high-count-rate performance of bismuth germanate

    International Nuclear Information System (INIS)

    Koehler, P.E.; Wender, S.A.; Kapustinsky, J.S.

    1985-01-01

    Several methods for improving the energy resolution of bismuth germanate (BGO) have been investigated. It is shown that some of these methods resulted in a substantial improvement in the energy resolution. In addition, a method to improve the performance of BGO at high counting rates has been systematically studied. The results of this study are presented and discussed

  16. Equivalence of truncated count mixture distributions and mixtures of truncated count distributions.

    Science.gov (United States)

    Böhning, Dankmar; Kuhnert, Ronny

    2006-12-01

    This article is about modeling count data with zero truncation. A parametric count density family is considered. The truncated mixture of densities from this family is different from the mixture of truncated densities from the same family. Whereas the former model is more natural to formulate and to interpret, the latter model is theoretically easier to treat. It is shown that for any mixing distribution leading to a truncated mixture, a (usually different) mixing distribution can be found so that the associated mixture of truncated densities equals the truncated mixture, and vice versa. This implies that the likelihood surfaces for both situations agree, and in this sense both models are equivalent. Zero-truncated count data models are used frequently in the capture-recapture setting to estimate population size, and it can be shown that the two Horvitz-Thompson estimators, associated with the two models, agree. In particular, it is possible to achieve strong results for mixtures of truncated Poisson densities, including reliable, global construction of the unique NPMLE (nonparametric maximum likelihood estimator) of the mixing distribution, implying a unique estimator for the population size. The benefit of these results lies in the fact that it is valid to work with the mixture of truncated count densities, which is less appealing for the practitioner but theoretically easier. Mixtures of truncated count densities form a convex linear model, for which a developed theory exists, including global maximum likelihood theory as well as algorithmic approaches. Once the problem has been solved in this class, it might readily be transformed back to the original problem by means of an explicitly given mapping. Applications of these ideas are given, particularly in the case of the truncated Poisson family.

  17. A 500-MHz x-ray counting system with a silicon avalanche photodiode

    International Nuclear Information System (INIS)

    Kishimoto, Shunji

    2009-01-01

    In the present measurements using a Si-APD X-ray detector and a 500-MHz counting system, the maximum output rate of 3.3x10 8 s -1 was achieved for 8-keV X-rays in beamline BL-14A of the Photon Factory. A small Si-APD of 4-pF electric capacity was used as the detector device in order to output a pulse of a width shorter than 2 ns on the baseline. For processing the fast pulses, the discriminator and the scaler having a throughput of >500 MHz, were prepared. Since the acceleration frequency at the PF ring was 500.1 MHz and the empty-bunch spacing was 12/312 bunches per circumference, the expected maximum rate was 4.8x10 8s-1 according to the counting model for a pulsed photon source. The reason why the present system did not reach the expected value was the baseline shift at the amplifier outputs. The rise of +0.2 V was observed at a discriminator output of 3.3x10 8 s -1 , while the pulse height was lower than 0.2 V. The baseline shift was caused by an AC coupling circuit in the amplifier. If a DC coupling circuit can be used for the amplifier, instead of the AC coupling circuit, or an active adjustment to compensate the baseline shift is installed, the counting system will show an ideal response. Although the present system including NIM modules was not so compact, we would like to develop a new fast-counting circuit for a Si-APD array detector of more than 100 channels of small pixels, in near future. (author)

  18. Heterogeneous counting on filter support media

    International Nuclear Information System (INIS)

    Long, E.; Kohler, V.; Kelly, M.J.

    1976-01-01

    Many investigators in the biomedical research area have used filter paper as the support for radioactive samples. This means that a heterogeneous counting of sample sometimes results. The count rate of a sample on a filter will be affected by positioning, degree of dryness, sample application procedure, the type of filter, and the type of cocktail used. Positioning of the filter (up or down) in the counting vial can cause a variation of 35% or more when counting tritiated samples on filter paper. Samples of varying degrees of dryness when added to the counting cocktail can cause nonreproducible counts if handled improperly. Count rates starting at 2400 CPM initially can become 10,000 CPM in 24 hours for 3 H-DNA (deoxyribonucleic acid) samples dried on standard cellulose acetate membrane filters. Data on cellulose nitrate filters show a similar trend. Sample application procedures in which the sample is applied to the filter in a small spot or on a large amount of the surface area can cause nonreproducible or very low counting rates. A tritiated DNA sample, when applied topically, gives a count rate of 4,000 CPM. When the sample is spread over the whole filter, 13,400 CPM are obtained with a much better coefficient of variation (5% versus 20%). Adding protein carrier (bovine serum albumin-BSA) to the sample to trap more of the tritiated DNA on the filter during the filtration process causes a serious beta absorption problem. Count rates which are one-fourth the count rate applied to the filter are obtained on calibrated runs. Many of the problems encountered can be alleviated by a proper choice of filter and the use of a liquid scintillation cocktail which dissolves the filter. Filter-Solv has been used to dissolve cellulose nitrate filters and filters which are a combination of cellulose nitrate and cellulose acetate. Count rates obtained for these dissolved samples are very reproducible and highly efficient

  19. High Resolution Gamma Ray Spectroscopy at MHz Counting Rates With LaBr3 Scintillators for Fusion Plasma Applications

    Science.gov (United States)

    Nocente, M.; Tardocchi, M.; Olariu, A.; Olariu, S.; Pereira, R. C.; Chugunov, I. N.; Fernandes, A.; Gin, D. B.; Grosso, G.; Kiptily, V. G.; Neto, A.; Shevelev, A. E.; Silva, M.; Sousa, J.; Gorini, G.

    2013-04-01

    High resolution γ-ray spectroscopy measurements at MHz counting rates were carried out at nuclear accelerators, combining a LaBr 3(Ce) detector with dedicated hardware and software solutions based on digitization and off-line analysis. Spectra were measured at counting rates up to 4 MHz, with little or no degradation of the energy resolution, adopting a pile up rejection algorithm. The reported results represent a step forward towards the final goal of high resolution γ-ray spectroscopy measurements on a burning plasma device.

  20. Correction for decay during counting in gamma spectrometry

    International Nuclear Information System (INIS)

    Nir-El, Y.

    2013-01-01

    A basic result in gamma spectrometry is the count rate of a relevant peak. Correction for decay during counting and expressing the count rate at the beginning of the measurement can be done by a multiplicative factor that is derived from integrating the count rate over time. The counting time substituted in this factor must be the live time, whereas the use of the real-time is an error that underestimates the count rate by about the dead-time (DT) (in percentage). This error of underestimation of the count rate is corroborated in the measurement of a nuclide with a high DT. The present methodology is not applicable in systems that include a zero DT correction function. (authors)

  1. Maximum inflation of the type 1 error rate when sample size and allocation rate are adapted in a pre-planned interim look.

    Science.gov (United States)

    Graf, Alexandra C; Bauer, Peter

    2011-06-30

    We calculate the maximum type 1 error rate of the pre-planned conventional fixed sample size test for comparing the means of independent normal distributions (with common known variance) which can be yielded when sample size and allocation rate to the treatment arms can be modified in an interim analysis. Thereby it is assumed that the experimenter fully exploits knowledge of the unblinded interim estimates of the treatment effects in order to maximize the conditional type 1 error rate. The 'worst-case' strategies require knowledge of the unknown common treatment effect under the null hypothesis. Although this is a rather hypothetical scenario it may be approached in practice when using a standard control treatment for which precise estimates are available from historical data. The maximum inflation of the type 1 error rate is substantially larger than derived by Proschan and Hunsberger (Biometrics 1995; 51:1315-1324) for design modifications applying balanced samples before and after the interim analysis. Corresponding upper limits for the maximum type 1 error rate are calculated for a number of situations arising from practical considerations (e.g. restricting the maximum sample size, not allowing sample size to decrease, allowing only increase in the sample size in the experimental treatment). The application is discussed for a motivating example. Copyright © 2011 John Wiley & Sons, Ltd.

  2. Ultrafast time measurements by time-correlated single photon counting coupled with superconducting single photon detector

    Energy Technology Data Exchange (ETDEWEB)

    Shcheslavskiy, V., E-mail: vis@becker-hickl.de; Becker, W. [Becker & Hickl GmbH, Nahmitzer Damm 30, 12277 Berlin (Germany); Morozov, P.; Divochiy, A. [Scontel, Rossolimo St., 5/22-1, Moscow 119021 (Russian Federation); Vakhtomin, Yu. [Scontel, Rossolimo St., 5/22-1, Moscow 119021 (Russian Federation); Moscow State Pedagogical University, 1/1 M. Pirogovskaya St., Moscow 119991 (Russian Federation); Smirnov, K. [Scontel, Rossolimo St., 5/22-1, Moscow 119021 (Russian Federation); Moscow State Pedagogical University, 1/1 M. Pirogovskaya St., Moscow 119991 (Russian Federation); National Research University Higher School of Economics, 20 Myasnitskaya St., Moscow 101000 (Russian Federation)

    2016-05-15

    Time resolution is one of the main characteristics of the single photon detectors besides quantum efficiency and dark count rate. We demonstrate here an ultrafast time-correlated single photon counting (TCSPC) setup consisting of a newly developed single photon counting board SPC-150NX and a superconducting NbN single photon detector with a sensitive area of 7 × 7 μm. The combination delivers a record instrument response function with a full width at half maximum of 17.8 ps and system quantum efficiency ∼15% at wavelength of 1560 nm. A calculation of the root mean square value of the timing jitter for channels with counts more than 1% of the peak value yielded about 7.6 ps. The setup has also good timing stability of the detector–TCSPC board.

  3. Low-Noise Free-Running High-Rate Photon-Counting for Space Communication and Ranging

    Science.gov (United States)

    Lu, Wei; Krainak, Michael A.; Yang, Guan; Sun, Xiaoli; Merritt, Scott

    2016-01-01

    We present performance data for low-noise free-running high-rate photon counting method for space optical communication and ranging. NASA GSFC is testing the performance of two types of novel photon-counting detectors 1) a 2x8 mercury cadmium telluride (HgCdTe) avalanche array made by DRS Inc., and a 2) a commercial 2880-element silicon avalanche photodiode (APD) array. We successfully measured real-time communication performance using both the 2 detected-photon threshold and logic AND-gate coincidence methods. Use of these methods allows mitigation of dark count, after-pulsing and background noise effects without using other method of Time Gating The HgCdTe APD array routinely demonstrated very high photon detection efficiencies (50) at near infrared wavelength. The commercial silicon APD array exhibited a fast output with rise times of 300 ps and pulse widths of 600 ps. On-chip individually filtered signals from the entire array were multiplexed onto a single fast output. NASA GSFC has tested both detectors for their potential application for space communications and ranging. We developed and compare their performances using both the 2 detected photon threshold and coincidence methods.

  4. Can simple mobile phone applications provide reliable counts of respiratory rates in sick infants and children? An initial evaluation of three new applications.

    Science.gov (United States)

    Black, James; Gerdtz, Marie; Nicholson, Pat; Crellin, Dianne; Browning, Laura; Simpson, Julie; Bell, Lauren; Santamaria, Nick

    2015-05-01

    Respiratory rate is an important sign that is commonly either not recorded or recorded incorrectly. Mobile phone ownership is increasing even in resource-poor settings. Phone applications may improve the accuracy and ease of counting of respiratory rates. The study assessed the reliability and initial users' impressions of four mobile phone respiratory timer approaches, compared to a 60-second count by the same participants. Three mobile applications (applying four different counting approaches plus a standard 60-second count) were created using the Java Mobile Edition and tested on Nokia C1-01 phones. Apart from the 60-second timer application, the others included a counter based on the time for ten breaths, and three based on the time interval between breaths ('Once-per-Breath', in which the user presses for each breath and the application calculates the rate after 10 or 20 breaths, or after 60s). Nursing and physiotherapy students used the applications to count respiratory rates in a set of brief video recordings of children with different respiratory illnesses. Limits of agreement (compared to the same participant's standard 60-second count), intra-class correlation coefficients and standard errors of measurement were calculated to compare the reliability of the four approaches, and a usability questionnaire was completed by the participants. There was considerable variation in the counts, with large components of the variation related to the participants and the videos, as well as the methods. None of the methods was entirely reliable, with no limits of agreement better than -10 to +9 breaths/min. Some of the methods were superior to the others, with ICCs from 0.24 to 0.92. By ICC the Once-per-Breath 60-second count and the Once-per-Breath 20-breath count were the most consistent, better even than the 60-second count by the participants. The 10-breath approaches performed least well. Users' initial impressions were positive, with little difference between the

  5. A relationship between salivary flow rates and Candida counts in patients with xerostomia.

    Science.gov (United States)

    Nadig, Suchetha Devendrappa; Ashwathappa, Deepak Timmasandra; Manjunath, Muniraju; Krishna, Sowmya; Annaji, Araleri Gopalkrishna; Shivaprakash, Praveen Kunigal

    2017-01-01

    Most of the adult population is colonized by Candida in their oral cavity. The process of colonization depends on several factors, including the interaction between Candida and salivary proteins. Therefore, salivary gland hypofunction may alter the oral microbiota and increase the risk for opportunistic infections, such as candidiasis. Hence, it is necessary to evaluate the relationship between salivary flow rates (SFRs) and Candida colony counts in the saliva of patients with xerostomia. This study aims to determine and evaluate the relationship between SFRs and Candida colony forming units (CFUs) in patients with xerostomia. This study was a descriptive study. The study participants were taken from the patients attending outpatient department in a private dental college. Fifty patients, who reported xerostomia in a questionnaire of the symptoms of xerostomia, were selected. Chewing stimulated whole saliva samples were collected from them and their SFRs were assessed. Saliva samples were inoculated in the Sabouraud dextrose agar culture media for 24-48 h, and Candida CFUs were counted. Chi-squared test was used to analyze the data. There was a significant inverse relationship between salivary flow and candida CFUs count when patients with high colony counts were analyzed (cutoff point of 400 or greater CFU/mL). Females had less SFR than males. Most of the patients who had hyposalivation were taking medication for the underlying systemic diseases. Candida albicans was the most frequent species. There was a significantly negative correlation between SFRs and Candida CFUs in the patients with xerostomia.

  6. Conifers in cold environments synchronize maximum growth rate of tree-ring formation with day length.

    Science.gov (United States)

    Rossi, Sergio; Deslauriers, Annie; Anfodillo, Tommaso; Morin, Hubert; Saracino, Antonio; Motta, Renzo; Borghetti, Marco

    2006-01-01

    Intra-annual radial growth rates and durations in trees are reported to differ greatly in relation to species, site and environmental conditions. However, very similar dynamics of cambial activity and wood formation are observed in temperate and boreal zones. Here, we compared weekly xylem cell production and variation in stem circumference in the main northern hemisphere conifer species (genera Picea, Pinus, Abies and Larix) from 1996 to 2003. Dynamics of radial growth were modeled with a Gompertz function, defining the upper asymptote (A), x-axis placement (beta) and rate of change (kappa). A strong linear relationship was found between the constants beta and kappa for both types of analysis. The slope of the linear regression, which corresponds to the time at which maximum growth rate occurred, appeared to converge towards the summer solstice. The maximum growth rate occurred around the time of maximum day length, and not during the warmest period of the year as previously suggested. The achievements of photoperiod could act as a growth constraint or a limit after which the rate of tree-ring formation tends to decrease, thus allowing plants to safely complete secondary cell wall lignification before winter.

  7. Maximum Acceptable Vibrato Excursion as a Function of Vibrato Rate in Musicians and Non-musicians

    DEFF Research Database (Denmark)

    Vatti, Marianna; Santurette, Sébastien; Pontoppidan, Niels H.

    2014-01-01

    and, in most listeners, exhibited a peak at medium vibrato rates (5–7 Hz). Large across-subject variability was observed, and no significant effect of musical experience was found. Overall, most listeners were not solely sensitive to the vibrato excursion and there was a listener-dependent rate...... for which larger vibrato excursions were favored. The observed interaction between maximum excursion thresholds and vibrato rate may be due to the listeners’ judgments relying on cues provided by the rate of frequency changes (RFC) rather than excursion per se. Further studies are needed to evaluate......Human vibrato is mainly characterized by two parameters: vibrato extent and vibrato rate. These parameters have been found to exhibit an interaction both in physical recordings of singers’ voices and in listener’s preference ratings. This study was concerned with the way in which the maximum...

  8. The Bayesian count rate probability distribution in measurement of ionizing radiation by use of a ratemeter

    Energy Technology Data Exchange (ETDEWEB)

    Weise, K.

    2004-06-01

    Recent metrological developments concerning measurement uncertainty, founded on Bayesian statistics, give rise to a revision of several parts of the DIN 25482 and ISO 11929 standard series. These series stipulate detection limits and decision thresholds for ionizing-radiation measurements. Part 3 and, respectively, part 4 of them deal with measurements by use of linear-scale analogue ratemeters. A normal frequency distribution of the momentary ratemeter indication for a fixed count rate value is assumed. The actual distribution, which is first calculated numerically by solving an integral equation, differs, however, considerably from the normal distribution although this one represents an approximation of it for sufficiently large values of the count rate to be measured. As is shown, this similarly holds true for the Bayesian probability distribution of the count rate for sufficiently large given measured values indicated by the ratemeter. This distribution follows from the first one mentioned by means of the Bayes theorem. Its expectation value and variance are needed for the standards to be revised on the basis of Bayesian statistics. Simple expressions are given by the present standards for estimating these parameters and for calculating the detection limit and the decision threshold. As is also shown, the same expressions can similarly be used as sufficient approximations by the revised standards if, roughly, the present indicated value exceeds the reciprocal ratemeter relaxation time constant. (orig.)

  9. LASER: A Maximum Likelihood Toolkit for Detecting Temporal Shifts in Diversification Rates From Molecular Phylogenies

    Directory of Open Access Journals (Sweden)

    Daniel L. Rabosky

    2006-01-01

    Full Text Available Rates of species origination and extinction can vary over time during evolutionary radiations, and it is possible to reconstruct the history of diversification using molecular phylogenies of extant taxa only. Maximum likelihood methods provide a useful framework for inferring temporal variation in diversification rates. LASER is a package for the R programming environment that implements maximum likelihood methods based on the birth-death process to test whether diversification rates have changed over time. LASER contrasts the likelihood of phylogenetic data under models where diversification rates have changed over time to alternative models where rates have remained constant over time. Major strengths of the package include the ability to detect temporal increases in diversification rates and the inference of diversification parameters under multiple rate-variable models of diversification. The program and associated documentation are freely available from the R package archive at http://cran.r-project.org.

  10. The study of error for analysis in dynamic image from the error of count rates in Nal (Tl) scintillation camera

    International Nuclear Information System (INIS)

    Oh, Joo Young; Kang, Chun Goo; Kim, Jung Yul; Oh, Ki Baek; Kim, Jae Sam; Park, Hoon Hee

    2013-01-01

    This study is aimed to evaluate the effect of T 1/2 upon count rates in the analysis of dynamic scan using NaI (Tl) scintillation camera, and suggest a new quality control method with this effects. We producted a point source with '9 9m TcO 4 - of 18.5 to 185 MBq in the 2 mL syringes, and acquired 30 frames of dynamic images with 10 to 60 seconds each using Infinia gamma camera (GE, USA). In the second experiment, 90 frames of dynamic images were acquired from 74 MBq point source by 5 gamma cameras (Infinia 2, Forte 2, Argus 1). There were not significant differences in average count rates of the sources with 18.5 to 92.5 MBq in the analysis of 10 to 60 seconds/frame with 10 seconds interval in the first experiment (p>0.05). But there were significantly low average count rates with the sources over 111 MBq activity at 60 seconds/frame (p<0.01). According to the second analysis results of linear regression by count rates of 5 gamma cameras those were acquired during 90 minutes, counting efficiency of fourth gamma camera was most low as 0.0064%, and gradient and coefficient of variation was high as 0.0042 and 0.229 each. We could not find abnormal fluctuation in χ 2 test with count rates (p>0.02), and we could find the homogeneity of variance in Levene's F-test among the gamma cameras (p>0.05). At the correlation analysis, there was only correlation between counting efficiency and gradient as significant negative correlation (r=-0.90, p<0.05). Lastly, according to the results of calculation of T 1/2 error from change of gradient with -0.25% to +0.25%, if T 1/2 is relatively long, or gradient is high, the error increase relationally. When estimate the value of 4th camera which has highest gradient from the above mentioned result, we could not see T 1/2 error within 60 minutes at that value. In conclusion, it is necessary for the scintillation gamma camera in medical field to manage hard for the quality of radiation measurement. Especially, we found a

  11. Maximum entropy production rate in quantum thermodynamics

    Energy Technology Data Exchange (ETDEWEB)

    Beretta, Gian Paolo, E-mail: beretta@ing.unibs.i [Universita di Brescia, via Branze 38, 25123 Brescia (Italy)

    2010-06-01

    In the framework of the recent quest for well-behaved nonlinear extensions of the traditional Schroedinger-von Neumann unitary dynamics that could provide fundamental explanations of recent experimental evidence of loss of quantum coherence at the microscopic level, a recent paper [Gheorghiu-Svirschevski 2001 Phys. Rev. A 63 054102] reproposes the nonlinear equation of motion proposed by the present author [see Beretta G P 1987 Found. Phys. 17 365 and references therein] for quantum (thermo)dynamics of a single isolated indivisible constituent system, such as a single particle, qubit, qudit, spin or atomic system, or a Bose-Einstein or Fermi-Dirac field. As already proved, such nonlinear dynamics entails a fundamental unifying microscopic proof and extension of Onsager's reciprocity and Callen's fluctuation-dissipation relations to all nonequilibrium states, close and far from thermodynamic equilibrium. In this paper we propose a brief but self-contained review of the main results already proved, including the explicit geometrical construction of the equation of motion from the steepest-entropy-ascent ansatz and its exact mathematical and conceptual equivalence with the maximal-entropy-generation variational-principle formulation presented in Gheorghiu-Svirschevski S 2001 Phys. Rev. A 63 022105. Moreover, we show how it can be extended to the case of a composite system to obtain the general form of the equation of motion, consistent with the demanding requirements of strong separability and of compatibility with general thermodynamics principles. The irreversible term in the equation of motion describes the spontaneous attraction of the state operator in the direction of steepest entropy ascent, thus implementing the maximum entropy production principle in quantum theory. The time rate at which the path of steepest entropy ascent is followed has so far been left unspecified. As a step towards the identification of such rate, here we propose a possible

  12. Association between success rate and citation count of studies of radiofrequency catheter ablation for atrial fibrillation: possible evidence of citation bias.

    Science.gov (United States)

    Perino, Alexander C; Hoang, Donald D; Holmes, Tyson H; Santangeli, Pasquale; Heidenreich, Paul A; Perez, Marco V; Wang, Paul J; Turakhia, Mintu P

    2014-09-01

    The preferential citation of studies with the highest success rates could exaggerate perceived effectiveness, particularly for treatments with widely varying published success rates such as radiofrequency catheter ablation for atrial fibrillation. We systematically identified observational studies and clinical trials of radiofrequency catheter ablation of atrial fibrillation between 1990 and 2012. Generalized Poisson regression was used to estimate association between study success rate and total citation count, adjusting for sample size, journal impact factor, time since publication, study design, and whether first or last author was a consensus-defined pre-eminent expert. We identified 174 articles meeting our inclusion criteria (36 289 subjects). After adjustment only for time since publication, a 10-point increase above the mean in pooled reported success rates was associated with a 17.8% increase in citation count at 5 years postpublication (95% confidence interval, 7.1-28.4%; Pcitation count; 95% confidence interval, 7.6-29.6%; Pcitation count, which may indicate citation bias. To readers of the literature, radiofrequency catheter ablation of atrial fibrillation could be perceived to be more effective than the data supports. These findings may have implications for a wide variety of novel cardiovascular therapies. © 2014 American Heart Association, Inc.

  13. The Dependence of Tropical Cyclone Count and Size on Rotation Rate

    Science.gov (United States)

    Chavas, D. R.; Reed, K. A.

    2017-12-01

    Both theory and idealized equilibrium modeling studies indicate that tropical cyclone size decreases with background rotation rate. In contrast, in real-world observations size tends to increase with latitude. Here we seek to resolve this apparent contradiction via a set of reduced-complexity global aquaplanet simulations with varying planetary rotation rates using the NCAR Community Atmosphere Model 5. The latitudinal distribution of both storm count and size are found to vary markedly with rotation rate, yielding insight into the dynamical constraints on tropical cyclone activity on a rotating planet. Moreover, storm size is found to vary non-monotonically with latitude, indicating that non-equilibrium effects are crucial to the life-cycle evolution of size in nature. Results are then compared to experiments in idealized, time-dependent limited-area modeling simulations using CM1 in axisymmetric and three-dimensional geometry. Taken together, this hierarchy of models is used to quantify the role of equilibrium versus transient controls on storm size and the relevance of each to real storms in nature.

  14. A relationship between salivary flow rates and Candida counts in patients with xerostomia

    OpenAIRE

    Nadig, Suchetha Devendrappa; Ashwathappa, Deepak Timmasandra; Manjunath, Muniraju; Krishna, Sowmya; Annaji, Araleri Gopalkrishna; Shivaprakash, Praveen Kunigal

    2017-01-01

    Context: Most of the adult population is colonized by Candida in their oral cavity. The process of colonization depends on several factors, including the interaction between Candida and salivary proteins. Therefore, salivary gland hypofunction may alter the oral microbiota and increase the risk for opportunistic infections, such as candidiasis. Hence, it is necessary to evaluate the relationship between salivary flow rates (SFRs) and Candida colony counts in the saliva of patients with xerost...

  15. Time and position resolution of high granularity, high counting rate MRPC for the inner zone of the CBM-TOF wall

    CERN Document Server

    Petriş, M.

    2016-09-13

    Multi-gap RPC prototypes with readout on a multi-strip electrode were developed for the small polar angle region of the CBM-TOF subdetector, the most demanding zone in terms of granularity and counting rate. The prototypes are based on low resistivity ($\\sim$10$^{10}$ $\\Omega$cm) glass electrodes for performing in high counting rate environment. The strip width/pitch size was chosen such to fulfill the impedance matching with the front-end electronics and the granularity requirements of the innermost zone of the CBM-TOF wall. The in-beam tests using secondary particles produced in heavy ion collisions on a Pb target at SIS18 - GSI Darmstadt and SPS - CERN were focused on the performance of the prototype in conditions similar to the ones expected at SIS100/FAIR. An efficiency larger than 98\\% and a system time resolution in the order of 70~-~80~ps were obtained in high counting rate and high multiplicity environment.

  16. Construction and Test of Muon Drift Tube Chambers for High Counting Rates

    CERN Document Server

    Schwegler, Philipp; Dubbert, Jörg

    2010-01-01

    Since the start of operation of the Large Hadron Collider (LHC) at CERN on 20 November 2009, the instantaneous luminosity is steadily increasing. The muon spectrometer of the ATLAS detector at the LHC is instrumented with trigger and precision tracking chambers in a toroidal magnetic field. Monitored Drift-Tube (MDT) chambers are employed as precision tracking chambers, complemented by Cathode Strip Chambers (CSC) in the very forward region where the background counting rate due to neutrons and γ's produced in shielding material and detector components is too high for the MDT chambers. After several upgrades of the CERN accelerator system over the coming decade, the instantaneous luminosity is expected to be raised to about five times the LHC design luminosity. This necessitates replacement of the muon chambers in the regions with the highest background radiation rates in the so-called Small Wheels, which constitute the innermost layers of the muon spectrometer end-caps, by new detectors with higher rate cap...

  17. Exponential decay and exponential recovery of modal gains in high count rate channel electron multipliers

    International Nuclear Information System (INIS)

    Hahn, S.F.; Burch, J.L.

    1980-01-01

    A series of data on high count rate channel electron multipliers revealed an initial drop and subsequent recovery of gains in exponential fashion. The FWHM of the pulse height distribution at the initial stage of testing can be used as a good criterion for the selection of operating bias voltage of the channel electron multiplier

  18. 7 CFR 4290.845 - Maximum rate of amortization on Loans and Debt Securities.

    Science.gov (United States)

    2010-01-01

    ...) RURAL BUSINESS-COOPERATIVE SERVICE AND RURAL UTILITIES SERVICE, DEPARTMENT OF AGRICULTURE RURAL BUSINESS INVESTMENT COMPANY (âRBICâ) PROGRAM Financing of Enterprises by RBICs Structuring Rbic Financing of Eligible Enterprises-Types of Financings § 4290.845 Maximum rate of amortization on Loans and Debt Securities. The...

  19. Investigation of the behaviour of both dead time and observed counting rates of He-3 gas filled neutron detector

    Energy Technology Data Exchange (ETDEWEB)

    Adib, M.; Eid, Y.; Abdel Kawy, A.; Maayouf, R.M.A.; Shuriet, G.M.; Hamouda, I.

    1981-01-01

    The behaviour of the dead time of He-3 detector, operating at both the proportional and the corona discharge regions, is investigated as a function of the neutron reaction rate inside the detector. The applied experimental method makes use of the fluctuations, due to the detector dead time in the observed counting rates from Poisson's distribution. In order to check the validity of the experimental method used in the present work, the dead time of BF/sub 3/ neutron detectors with different efficiencies (due to different enrichement in B-10) were determined. It is shown that the observed counting rate from the He-3 detector operating at the proportional region for neutron reaction rates ranging from 8 x 10/sup 3/ to 2.5 x 10/sup 4/ reaction/sec decreases with the increase of the neutron reaction rate. Such behaviour was not observed when operating the He-3 detector at the corona discharge region.

  20. Investigation of the behaviour of both dead time and observed counting rates of He-3 gas filled neutron detector

    International Nuclear Information System (INIS)

    Adib, M.; Eid, Y.; Abdel Kawy, A.; Maayouf, R.M.A.; Shuriet, G.M.; Hamouda, I.

    1981-01-01

    The behaviour of the dead time of He-3 detector, operating at both the proportional and the corona discharge regions, is investigated as a function of the neutron reaction rate inside the detector. The applied experimental method makes use of the fluctuations, due to the detector dead time in the observed counting rates from Poisson's distribution. In order to check the validity of the experimental method used in the present work, the dead time of BF 3 neutron detectors with different efficiencies (due to different enrichement in B-10) were determined. It is shown that the observed counting rate from the He-3 detector operating at the proportional region for neutron reaction rates ranging from 8 x 10 3 to 2.5 x 10 4 reaction/sec decreases with the increase of the neutron reaction rate. Such behaviour was not observed when operating the He-3 detector at the corona discharge region. (orig.) [de

  1. A generalization of the preset count moving average algorithm for digital rate meters

    International Nuclear Information System (INIS)

    Arandjelovic, Vojislav; Koturovic, Aleksandar; Vukanovic, Radomir

    2002-01-01

    A generalized definition of the preset count moving average algorithm for digital rate meters has been introduced. The algorithm is based on the knowledge of time intervals between successive pulses in random-pulse sequences. The steady state and transient regimes of the algorithm have been characterized. A measure for statistical fluctuations of the successive measurement results has been introduced. The versatility of the generalized algorithm makes it suitable for application in the design of the software of modern measuring/control digital systems

  2. Tower counts

    Science.gov (United States)

    Woody, Carol Ann; Johnson, D.H.; Shrier, Brianna M.; O'Neal, Jennifer S.; Knutzen, John A.; Augerot, Xanthippe; O'Neal, Thomas A.; Pearsons, Todd N.

    2007-01-01

    Counting towers provide an accurate, low-cost, low-maintenance, low-technology, and easily mobilized escapement estimation program compared to other methods (e.g., weirs, hydroacoustics, mark-recapture, and aerial surveys) (Thompson 1962; Siebel 1967; Cousens et al. 1982; Symons and Waldichuk 1984; Anderson 2000; Alaska Department of Fish and Game 2003). Counting tower data has been found to be consistent with that of digital video counts (Edwards 2005). Counting towers do not interfere with natural fish migration patterns, nor are fish handled or stressed; however, their use is generally limited to clear rivers that meet specific site selection criteria. The data provided by counting tower sampling allow fishery managers to determine reproductive population size, estimate total return (escapement + catch) and its uncertainty, evaluate population productivity and trends, set harvest rates, determine spawning escapement goals, and forecast future returns (Alaska Department of Fish and Game 1974-2000 and 1975-2004). The number of spawning fish is determined by subtracting subsistence, sport-caught fish, and prespawn mortality from the total estimated escapement. The methods outlined in this protocol for tower counts can be used to provide reasonable estimates ( plus or minus 6%-10%) of reproductive salmon population size and run timing in clear rivers. 

  3. Growth and maximum size of tiger sharks (Galeocerdo cuvier) in Hawaii.

    Science.gov (United States)

    Meyer, Carl G; O'Malley, Joseph M; Papastamatiou, Yannis P; Dale, Jonathan J; Hutchinson, Melanie R; Anderson, James M; Royer, Mark A; Holland, Kim N

    2014-01-01

    Tiger sharks (Galecerdo cuvier) are apex predators characterized by their broad diet, large size and rapid growth. Tiger shark maximum size is typically between 380 & 450 cm Total Length (TL), with a few individuals reaching 550 cm TL, but the maximum size of tiger sharks in Hawaii waters remains uncertain. A previous study suggested tiger sharks grow rather slowly in Hawaii compared to other regions, but this may have been an artifact of the method used to estimate growth (unvalidated vertebral ring counts) compounded by small sample size and narrow size range. Since 1993, the University of Hawaii has conducted a research program aimed at elucidating tiger shark biology, and to date 420 tiger sharks have been tagged and 50 recaptured. All recaptures were from Hawaii except a single shark recaptured off Isla Jacques Cousteau (24°13'17″N 109°52'14″W), in the southern Gulf of California (minimum distance between tag and recapture sites  =  approximately 5,000 km), after 366 days at liberty (DAL). We used these empirical mark-recapture data to estimate growth rates and maximum size for tiger sharks in Hawaii. We found that tiger sharks in Hawaii grow twice as fast as previously thought, on average reaching 340 cm TL by age 5, and attaining a maximum size of 403 cm TL. Our model indicates the fastest growing individuals attain 400 cm TL by age 5, and the largest reach a maximum size of 444 cm TL. The largest shark captured during our study was 464 cm TL but individuals >450 cm TL were extremely rare (0.005% of sharks captured). We conclude that tiger shark growth rates and maximum sizes in Hawaii are generally consistent with those in other regions, and hypothesize that a broad diet may help them to achieve this rapid growth by maximizing prey consumption rates.

  4. The effect of event shape on centroiding in photon counting detectors

    International Nuclear Information System (INIS)

    Kawakami, Hajime; Bone, David; Fordham, John; Michel, Raul

    1994-01-01

    High resolution, CCD readout, photon counting detectors employ simple centroiding algorithms for defining the spatial position of each detected event. The accuracy of centroiding is very dependent upon a number of parameters including the profile, energy and width of the intensified event. In this paper, we provide an analysis of how the characteristics of an intensified event change as the input count rate increases and the consequent effect on centroiding. The changes in these parameters are applied in particular to the MIC photon counting detector developed at UCL for ground and space based astronomical applications. This detector has a maximum format of 3072x2304 pixels permitting its use in the highest resolution applications. Individual events, at light level from 5 to 1000k events/s over the detector area, were analysed. It was found that both the asymmetry and width of event profiles were strongly dependent upon the energy of the intensified event. The variation in profile then affected the centroiding accuracy leading to loss of resolution. These inaccuracies have been quantified for two different 3 CCD pixel centroiding algorithms and one 2 pixel algorithm. The results show that a maximum error of less than 0.05 CCD pixel occurs with the 3 pixel algorithms and 0.1 CCD pixel for the 2 pixel algorithm. An improvement is proposed by utilising straight pore MCPs in the intensifier and a 70 μm air gap in front of the CCD. ((orig.))

  5. Relationship of long-term highly active antiretroviral therapy on salivary flow rate and CD4 Count among HIV-infected patients.

    Science.gov (United States)

    Kumar, J Vijay; Baghirath, P Venkat; Naishadham, P Parameswar; Suneetha, Sujai; Suneetha, Lavanya; Sreedevi, P

    2015-01-01

    To determine if long-term highly active antiretroviral therapy (HAART) therapy alters salivary flow rate and also to compare its relation of CD4 count with unstimulated and stimulated whole saliva. A cross-sectional study was performed on 150 individuals divided into three groups. Group I (50 human immunodeficiency virus (HIV) seropositive patients, but not on HAART therapy), Group II (50 HIV-infected subjects and on HAART for less than 3 years called short-term HAART), Group III (50 HIV-infected subjects and on HAART for more than or equal to 3 years called long-term HAART). Spitting method proposed by Navazesh and Kumar was used for the measurement of unstimulated and stimulated salivary flow rate. Chi-square test and analysis of variance (ANOVA) were used for statistical analysis. The mean CD4 count was 424.78 ± 187.03, 497.82 ± 206.11 and 537.6 ± 264.00 in the respective groups. Majority of the patients in all the groups had a CD4 count between 401 and 600. Both unstimulated and stimulated whole salivary (UWS and SWS) flow rates in Group I was found to be significantly higher than in Group II (P flow rate between Group II and III subjects were also found to be statistically significant (P relationship in Group II (P flow rates of HIV-infected individuals who are on long-term HAART.

  6. Two dimensional localization of electrons and positrons under high counting rate

    International Nuclear Information System (INIS)

    Barbosa, A.F.; Anjos, J.C.; Sanchez-Hernandez, A.; Pepe, I.M.; Barros, N.

    1997-12-01

    The construction of two wire chambers for the experiment E831 at Fermilab is reported. Each chamber includes three wire planes - one anode and two orthogonal cathodes - in which the wires operate as independent proportional counters. One of the chambers is rotated with respect to the other, so that four position coordinates may be encoded for a charged particle crossing both chambers. Spatial resolution is determined by the wire pitch: 1 mm for cathodes, 2 mm for anodes. 320 electronic channels are involved in the detection system readout. Global counting rates in excess to 10 7 events per second have been measured, while the average electron-positron beam intensity may be as high as 3 x 10 7 events per second. (author)

  7. Relationship of blood and milk cell counts with mastitic pathogens in Murrah buffaloes

    Directory of Open Access Journals (Sweden)

    C. Singh

    2010-02-01

    Full Text Available The present study was undertaken to see the effect of mastitic pathogens on the blood and milk counts of Murrah buffaloes. Milk and blood samples were collected from 9 mastitic Murrah buffaloes. The total leucocyte Counts (TLC and Differential leucocyte counts (DLC in blood were within normal range and there was a non-significant change in blood counts irrespective of different mastitic pathogens. Normal milk quarter samples had significantly (P<0.01 less Somatic cell counts (SCC. Lymphocytes were significantly higher in normal milk samples, whereas infected samples had a significant increase (P<0.01 in milk neutrophils. S. aureus infected buffaloes had maximum milk SCC, followed by E. coli and S. agalactiae. Influx of neutrophils in the buffalo mammary gland was maximum for S. agalactiae, followed by E.cli and S. aureus. The study indicated that level of mastitis had no affect on blood counts but it influenced the milk SCC of normal quarters.

  8. Eutectic cell and nodule count as the quality factors of cast iron

    Directory of Open Access Journals (Sweden)

    E. Fraś

    2008-10-01

    Full Text Available In this work the predictions based on a theoretical analysis aimed at elucidating of eutectic cell count or nodule counts N wereexperimentally verified. The experimental work was focused on processing flake graphite and ductile iron under various inoculationconditions in order to achieve various physicochemical states of the experimental melts. In addition, plates of various wall thicknesses, s were cast and the resultant eutectic cell or nodule counts were established. Moreover, thermal analysis was used to find out the degree of maximum undercooling for the graphite eutectic, Tm. A relationship was found between the eutectic cell or nodule count and the maximum undercooling Tm.. In addition it was also found that N can be related to the wall thickness of plate shaped castings. Finally, the present work provides a rational for the effect of technological factors such as the melt chemistry, inoculation practice, and holding temperature and time on the resultant cell count or nodule count of cast iron. In particular, good agreement was found between the predictions of the theoretical analysis and the experimental data.

  9. Photon-Counting Arrays for Time-Resolved Imaging

    Directory of Open Access Journals (Sweden)

    I. Michel Antolovic

    2016-06-01

    Full Text Available The paper presents a camera comprising 512 × 128 pixels capable of single-photon detection and gating with a maximum frame rate of 156 kfps. The photon capture is performed through a gated single-photon avalanche diode that generates a digital pulse upon photon detection and through a digital one-bit counter. Gray levels are obtained through multiple counting and accumulation, while time-resolved imaging is achieved through a 4-ns gating window controlled with subnanosecond accuracy by a field-programmable gate array. The sensor, which is equipped with microlenses to enhance its effective fill factor, was electro-optically characterized in terms of sensitivity and uniformity. Several examples of capture of fast events are shown to demonstrate the suitability of the approach.

  10. Performance of a GM tube based environmental dose rate monitor operating in the Time-To-Count mode

    Energy Technology Data Exchange (ETDEWEB)

    Zickefoose, J.; Kulkarni, T.; Martinson, T.; Phillips, K.; Voelker, M. [Canberra Industries Inc. (United States)

    2015-07-01

    The events at the Fukushima Daiichi power plant in the aftermath of a natural disaster underline the importance of a large array of networked environmental monitors to cover areas around nuclear power plants. These monitors should meet a few basic criteria: have a uniform response over a wide range of gamma energies, have a uniform response over a wide range of incident angles, and have a large dynamic range. Many of these criteria are met if the probe is qualified to the international standard IEC 60532 (Radiation protection instrumentation - Installed dose rate meters, warning assemblies and monitors - X and gamma radiation of energy between 50 keV and 7 MeV), which specifically deals with energy response, angle of incidence, dynamic range, response time, and a number of environmental characteristics. EcoGamma is a dual GM tube environmental gamma radiation monitor designed specifically to meet the requirements of IEC 60532 and operate in the most extreme conditions. EcoGamma utilizes two energy compensated GM tubes operating with a Time-To-Count (TTC) collection algorithm. The TTC algorithm extends the lifetime and range of a GM tube significantly and allows the dual GM tube probe to achieve linearity over approximately 10 decades of gamma dose rate (from the Sv/hr range to 100 Sv/hr). In the TTC mode of operation, the GM tube is not maintained in a biased condition continuously. This is different from a traditional counting system where the GM tube is held at a constant bias continuously and the total number of strikes that the tube registers are counted. The traditional approach allows for good sensitivity, but does not lend itself to a long lifetime of the tube and is susceptible to linearity issues at high count rates. TTC on the other hand only biases the tube for short periods of time and in effect measures the time between events, which is statistically representative of the total strike rate. Since the tube is not continually biased, the life of the tube

  11. Performance of a GM tube based environmental dose rate monitor operating in the Time-To-Count mode

    International Nuclear Information System (INIS)

    Zickefoose, J.; Kulkarni, T.; Martinson, T.; Phillips, K.; Voelker, M.

    2015-01-01

    The events at the Fukushima Daiichi power plant in the aftermath of a natural disaster underline the importance of a large array of networked environmental monitors to cover areas around nuclear power plants. These monitors should meet a few basic criteria: have a uniform response over a wide range of gamma energies, have a uniform response over a wide range of incident angles, and have a large dynamic range. Many of these criteria are met if the probe is qualified to the international standard IEC 60532 (Radiation protection instrumentation - Installed dose rate meters, warning assemblies and monitors - X and gamma radiation of energy between 50 keV and 7 MeV), which specifically deals with energy response, angle of incidence, dynamic range, response time, and a number of environmental characteristics. EcoGamma is a dual GM tube environmental gamma radiation monitor designed specifically to meet the requirements of IEC 60532 and operate in the most extreme conditions. EcoGamma utilizes two energy compensated GM tubes operating with a Time-To-Count (TTC) collection algorithm. The TTC algorithm extends the lifetime and range of a GM tube significantly and allows the dual GM tube probe to achieve linearity over approximately 10 decades of gamma dose rate (from the Sv/hr range to 100 Sv/hr). In the TTC mode of operation, the GM tube is not maintained in a biased condition continuously. This is different from a traditional counting system where the GM tube is held at a constant bias continuously and the total number of strikes that the tube registers are counted. The traditional approach allows for good sensitivity, but does not lend itself to a long lifetime of the tube and is susceptible to linearity issues at high count rates. TTC on the other hand only biases the tube for short periods of time and in effect measures the time between events, which is statistically representative of the total strike rate. Since the tube is not continually biased, the life of the tube

  12. Effects of the thickness of gold deposited on a source backing film in the 4πβ-counting

    International Nuclear Information System (INIS)

    Miyahara, Hiroshi; Yoshida, Makoto; Watanabe, Tamaki

    1976-01-01

    A gold deposited VYNS film as a source backing in the 4πβ-counting has generally been used for reducing the absorption of β-rays. The thickness of the film with the gold is usually a few times thicker than the VYNS film itself. However, Because the appropriate thickness of gold has not yet been determined, the effects of gold thickness on electrical resistivity, plateau characteristics and β-ray counting efficiency were studied. 198 Au (960 keV), 60 Co(315 keV), 59 Fe(273 keV) and 95 Nb(160 keV), which were prepared as sources by the aluminium chloride treatment method, were used. Gold was evaporated under a deposition rate of 1 - 5 μg/cm 2 /min at a pressure less than 1 x 10 -5 Torr. Results show that the gold deposition on the side opposite the source after source preparation is essential. In this case, a maximum counting efficiency is obtained at the mean thickness of 2 μg/cm 2 . When gold is deposited only on the same side as the source, a maximum counting efficiency, which is less than that in the former case, is obtained at the mean thickness of 20 μg/cm 2 . (Evans, J.)

  13. Phase space representation of neutron monitor count rate and atmospheric electric field in relation to solar activity in cycles 21 and 22.

    Science.gov (United States)

    Silva, H G; Lopes, I

    Heliospheric modulation of galactic cosmic rays links solar cycle activity with neutron monitor count rate on earth. A less direct relation holds between neutron monitor count rate and atmospheric electric field because different atmospheric processes, including fluctuations in the ionosphere, are involved. Although a full quantitative model is still lacking, this link is supported by solid statistical evidence. Thus, a connection between the solar cycle activity and atmospheric electric field is expected. To gain a deeper insight into these relations, sunspot area (NOAA, USA), neutron monitor count rate (Climax, Colorado, USA), and atmospheric electric field (Lisbon, Portugal) are presented here in a phase space representation. The period considered covers two solar cycles (21, 22) and extends from 1978 to 1990. Two solar maxima were observed in this dataset, one in 1979 and another in 1989, as well as one solar minimum in 1986. Two main observations of the present study were: (1) similar short-term topological features of the phase space representations of the three variables, (2) a long-term phase space radius synchronization between the solar cycle activity, neutron monitor count rate, and potential gradient (confirmed by absolute correlation values above ~0.8). Finally, the methodology proposed here can be used for obtaining the relations between other atmospheric parameters (e.g., solar radiation) and solar cycle activity.

  14. New England observed and predicted August stream/river temperature maximum daily rate of change points

    Data.gov (United States)

    U.S. Environmental Protection Agency — The shapefile contains points with associated observed and predicted August stream/river temperature maximum negative rate of change in New England based on a...

  15. An Adaptive Smoother for Counting Measurements

    International Nuclear Information System (INIS)

    Kondrasovs Vladimir; Coulon Romain; Normand Stephane

    2013-06-01

    Counting measurements associated with nuclear instruments are tricky to carry out due to the stochastic process of the radioactivity. Indeed events counting have to be processed and filtered in order to display a stable count rate value and to allow variations monitoring in the measured activity. Smoothers (as the moving average) are adjusted by a time constant defined as a compromise between stability and response time. A new approach has been developed and consists in improving the response time while maintaining count rate stability. It uses the combination of a smoother together with a detection filter. A memory of counting data is processed to calculate several count rate estimates using several integration times. These estimates are then sorted into the memory from short to long integration times. A measurement position, in terms of integration time, is then chosen into this memory after a detection test. An inhomogeneity into the Poisson counting process is detected by comparison between current position estimate and the other estimates contained into the memory in respect with the associated statistical variance calculated with homogeneous assumption. The measurement position (historical time) and the ability to forget an obsolete data or to keep in memory a useful data are managed using the detection test result. The proposed smoother is then an adaptive and a learning algorithm allowing an optimization of the response time while maintaining measurement counting stability and converging efficiently to the best counting estimate after an effective change in activity. This algorithm has also the specificity to be low recursive and thus easily embedded into DSP electronics based on FPGA or micro-controllers meeting 'real life' time requirements. (authors)

  16. Fast imaging by photon counting application to long-baseline optical stellar interferometry

    International Nuclear Information System (INIS)

    Morel, Sebastien

    1998-01-01

    Image acquisition by photon counting in the visible spectrum with a high precision on photo-events dating is especially useful for ground-based observations. In the first part of this thesis, and after a review of several techniques for photon acquisition and processing, I introduce a new type of photon counting camera, noticeable for its high temporal resolution and its high maximum counting rate: the DELTA (Detector Enhancement by Linear-projections on Three Axes) camera. I describe the concept of this camera, and the engineering solutions (optics, electronics, computing) that could be used for its construction. The second part of my work regards fringe detection and tracking in ground-based and long- baseline optical stellar interferometry. After a statistical approach of the issue, I describe methods introducing a priori information in the data, in order to have a better detection efficiency. One of the proposed methods, using a priori information on the atmospheric piston, requires a precise photo-event dating, and therefore uses methods described in the first part. (author) [fr

  17. Analysis of reaction schemes using maximum rates of constituent steps

    Science.gov (United States)

    Motagamwala, Ali Hussain; Dumesic, James A.

    2016-01-01

    We show that the steady-state kinetics of a chemical reaction can be analyzed analytically in terms of proposed reaction schemes composed of series of steps with stoichiometric numbers equal to unity by calculating the maximum rates of the constituent steps, rmax,i, assuming that all of the remaining steps are quasi-equilibrated. Analytical expressions can be derived in terms of rmax,i to calculate degrees of rate control for each step to determine the extent to which each step controls the rate of the overall stoichiometric reaction. The values of rmax,i can be used to predict the rate of the overall stoichiometric reaction, making it possible to estimate the observed reaction kinetics. This approach can be used for catalytic reactions to identify transition states and adsorbed species that are important in controlling catalyst performance, such that detailed calculations using electronic structure calculations (e.g., density functional theory) can be carried out for these species, whereas more approximate methods (e.g., scaling relations) are used for the remaining species. This approach to assess the feasibility of proposed reaction schemes is exact for reaction schemes where the stoichiometric coefficients of the constituent steps are equal to unity and the most abundant adsorbed species are in quasi-equilibrium with the gas phase and can be used in an approximate manner to probe the performance of more general reaction schemes, followed by more detailed analyses using full microkinetic models to determine the surface coverages by adsorbed species and the degrees of rate control of the elementary steps. PMID:27162366

  18. Low-noise low-jitter 32-pixels CMOS single-photon avalanche diodes array for single-photon counting from 300 nm to 900 nm

    Energy Technology Data Exchange (ETDEWEB)

    Scarcella, Carmelo; Tosi, Alberto, E-mail: alberto.tosi@polimi.it; Villa, Federica; Tisa, Simone; Zappa, Franco [Politecnico di Milano, Dipartimento di Elettronica, Informazione e Bioingegneria, Piazza Leonardo da Vinci 32, I-20133 Milano (Italy)

    2013-12-15

    We developed a single-photon counting multichannel detection system, based on a monolithic linear array of 32 CMOS SPADs (Complementary Metal-Oxide-Semiconductor Single-Photon Avalanche Diodes). All channels achieve a timing resolution of 100 ps (full-width at half maximum) and a photon detection efficiency of 50% at 400 nm. Dark count rate is very low even at room temperature, being about 125 counts/s for 50 μm active area diameter SPADs. Detection performance and microelectronic compactness of this CMOS SPAD array make it the best candidate for ultra-compact time-resolved spectrometers with single-photon sensitivity from 300 nm to 900 nm.

  19. Performance of Drift-Tube Detectors at High Counting Rates for High-Luminosity LHC Upgrades

    CERN Document Server

    Bittner, Bernhard; Kortner, Oliver; Kroha, Hubert; Manfredini, Alessandro; Nowak, Sebastian; Ott, Sebastian; Richter, Robert; Schwegler, Philipp; Zanzi, Daniele; Biebel, Otmar; Hertenberger, Ralf; Ruschke, Alexander; Zibell, Andre

    2016-01-01

    The performance of pressurized drift-tube detectors at very high background rates has been studied at the Gamma Irradiation Facility (GIF) at CERN and in an intense 20 MeV proton beam at the Munich Van-der-Graaf tandem accelerator for applications in large-area precision muon tracking at high-luminosity upgrades of the Large Hadron Collider (LHC). The ATLAS muon drifttube (MDT) chambers with 30 mm tube diameter have been designed to cope with and neutron background hit rates of up to 500 Hz/square cm. Background rates of up to 14 kHz/square cm are expected at LHC upgrades. The test results with standard MDT readout electronics show that the reduction of the drift-tube diameter to 15 mm, while leaving the operating parameters unchanged, vastly increases the rate capability well beyond the requirements. The development of new small-diameter muon drift-tube (sMDT) chambers for LHC upgrades is completed. Further improvements of tracking e?ciency and spatial resolution at high counting rates will be achieved with ...

  20. Regression analysis of mixed recurrent-event and panel-count data with additive rate models.

    Science.gov (United States)

    Zhu, Liang; Zhao, Hui; Sun, Jianguo; Leisenring, Wendy; Robison, Leslie L

    2015-03-01

    Event-history studies of recurrent events are often conducted in fields such as demography, epidemiology, medicine, and social sciences (Cook and Lawless, 2007, The Statistical Analysis of Recurrent Events. New York: Springer-Verlag; Zhao et al., 2011, Test 20, 1-42). For such analysis, two types of data have been extensively investigated: recurrent-event data and panel-count data. However, in practice, one may face a third type of data, mixed recurrent-event and panel-count data or mixed event-history data. Such data occur if some study subjects are monitored or observed continuously and thus provide recurrent-event data, while the others are observed only at discrete times and hence give only panel-count data. A more general situation is that each subject is observed continuously over certain time periods but only at discrete times over other time periods. There exists little literature on the analysis of such mixed data except that published by Zhu et al. (2013, Statistics in Medicine 32, 1954-1963). In this article, we consider the regression analysis of mixed data using the additive rate model and develop some estimating equation-based approaches to estimate the regression parameters of interest. Both finite sample and asymptotic properties of the resulting estimators are established, and the numerical studies suggest that the proposed methodology works well for practical situations. The approach is applied to a Childhood Cancer Survivor Study that motivated this study. © 2014, The International Biometric Society.

  1. Automated uranium analysis by delayed-neutron counting

    International Nuclear Information System (INIS)

    Kunzendorf, H.; Loevborg, L.; Christiansen, E.M.

    1980-10-01

    Automated uranium analysis by fission-induced delayed-neutron counting is described. A short description is given of the instrumentation including transfer system, process control, irradiation and counting sites, and computer operations. Characteristic parameters of the facility (sample preparations, background, and standards) are discussed. A sensitivity of 817 +- 22 counts per 10 -6 g U is found using irradiation, delay, and counting times of 20 s, 5 s, and 10 s, respectively. Presicion is generally less than 1% for normal geological samples. Critical level and detection limits for 7.5 g samples are 8 and 16 ppb, respectively. The importance of some physical and elemental interferences are outlined. Dead-time corrections of measured count rates are necessary and a polynomical expression is used for count rates up to 10 5 . The presence of rare earth elements is regarded as the most important elemental interference. A typical application is given and other areas of application are described. (auther)

  2. Influence of materials and counting-rate effects on 3He neutron spectrometry

    International Nuclear Information System (INIS)

    Evans, A.E.

    1984-01-01

    The high energy resolution of the Cuttler-Shalev 3 He neutron spectrometer causes spectral measurements with this instrument to be strongly susceptible to artifacts caused by the presence of scattering or absorbing materials in or near the detector or the source, and to false peaks generated by pileup coincidences of the rather long-risetime pulses from the detector. These effects are particularly important when pulse-height distributions vary over several orders of magnitude in count rate versus channel. A commercial pile-up elimination circuit greatly improves but does not eliminate the pileup problem. Previously reported spurious peaks in the pulse-height distributions from monoenergetic neutron sources have been determined to be due to the influence of the iron in the detector wall. 6 references, 9 figures

  3. Signal shaping and tail cancellation for gas proportional detectors at high counting rates

    International Nuclear Information System (INIS)

    Boie, R.A.; Hrisoho, A.T.; Rehak, P.

    1982-01-01

    A low noise, wide bandwidth preamplifier and signal processing filter were developed for high counting rate proportional counters. The filter consists of a seven pole Gaussian integrator with symmetrical weighting function and continuously variable shaping time, tausub(s), of 8-50 ns (fwhm) preceded by a second order pole/zero circuit which cancels the long (1/t) tails of the chamber signals. The preamplifier is an optimized common base input design with 2 ns rise time and an equivalent noise input charge < 2000 r.m.s. electrons, when connected to a chamber with 10 pF capacitance and at a filtering time, tausub(s), of 10 ns. (orig.)

  4. Development of bonded semiconductor device for high counting rate high efficiency photon detectors

    International Nuclear Information System (INIS)

    Kanno, Ikuo

    2008-01-01

    We are trying to decrease dose exposure in medical diagnosis by way of measuring the energy of X-rays. For this purpose, radiation detectors for X-ray energy measurement with high counting rate should be developed. Direct bonding of Si wafers was carried out to make a radiation detector, which had separated X-ray absorber and detector. The resistivity of bonding interface was estimated with the results of four-probe measurements and model calculations. Direct bonding of high resistivity p and n-Si wafers was also performed. The resistance of the pn bonded diode was 0.7 MΩ. The resistance should be increased in the future. (author)

  5. Variable sampling-time technique for improving count rate performance of scintillation detectors

    International Nuclear Information System (INIS)

    Tanaka, E.; Nohara, N.; Murayama, H.

    1979-01-01

    A new technique is presented to improve the count rate capability of a scintillation spectrometer or a position sensitive detector with minimum loss of resolution. The technique is based on the combination of pulse shortening and selective integration in which the integration period is not fixed but shortened by the arrival of the following pulse. Theoretical analysis of the degradation of the statiscal component of resolution is made for the proposed system with delay line pulse shortening, and the factor of resolution loss is formulated as a function of the input pulse rate. A new method is also presented for determining the statistical component of resolution separately from the non-statistical system resolution. Preliminary experiments with a NaI(Tl) detector have been carried out, the results of which are consistent with the theoretical prediction. However, due to the non-exponential scintillation decay of the NaI(Tl) crystal, a simple delay line clipping is not satisfactory, and an RC high-pass filter has been added, which results in further degradation of the statistical resolution. (Auth.)

  6. New England observed and predicted August stream/river temperature maximum positive daily rate of change points

    Data.gov (United States)

    U.S. Environmental Protection Agency — The shapefile contains points with associated observed and predicted August stream/river temperature maximum positive daily rate of change in New England based on a...

  7. New England observed and predicted July stream/river temperature maximum positive daily rate of change points

    Data.gov (United States)

    U.S. Environmental Protection Agency — The shapefile contains points with associated observed and predicted July stream/river temperature maximum positive daily rate of change in New England based on a...

  8. New England observed and predicted July maximum negative stream/river temperature daily rate of change points

    Data.gov (United States)

    U.S. Environmental Protection Agency — The shapefile contains points with associated observed and predicted July stream/river temperature maximum negative daily rate of change in New England based on a...

  9. Novel Photon-Counting Detectors for Free-Space Communication

    Science.gov (United States)

    Krainak, Michael A.; Yang, Guan; Sun, Xiaoli; Lu, Wei; Merritt, Scott; Beck, Jeff

    2016-01-01

    We present performance data for novel photon counting detectors for free space optical communication. NASA GSFC is testing the performance of three novel photon counting detectors 1) a 2x8 mercury cadmium telluride avalanche array made by DRS Inc. 2) a commercial 2880 silicon avalanche photodiode array and 3) a prototype resonant cavity silicon avalanche photodiode array. We will present and compare dark count, photon detection efficiency, wavelength response and communication performance data for these detectors. We discuss system wavelength trades and architectures for optimizing overall communication link sensitivity, data rate and cost performance. The HgCdTe APD array has photon detection efficiencies of greater than 50 were routinely demonstrated across 5 arrays, with one array reaching a maximum PDE of 70. High resolution pixel-surface spot scans were performed and the junction diameters of the diodes were measured. The junction diameter was decreased from 31 m to 25 m resulting in a 2x increase in e-APD gain from 470 on the 2010 array to 1100 on the array delivered to NASA GSFC. Mean single photon SNRs of over 12 were demonstrated at excess noise factors of 1.2-1.3.The commercial silicon APD array has a fast output with rise times of 300ps and pulse widths of 600ps. Received and filtered signals from the entire array are multiplexed onto this single fast output. The prototype resonant cavity silicon APD array is being developed for use at 1 micron wavelength.

  10. Comparing sensitivity and count rate performance of small-bore DOI-PET scanners by computer simulation

    International Nuclear Information System (INIS)

    Kobayashi, Tetsuya; Takahashi, Hisashi; Yamaya, Taiga; Murayama, Hideo; Kitamura, Keishi; Hasegawa, Tomoyuki; Suga, Mikio

    2006-01-01

    In the development of a dedicated small-bore DOI-PET scanner for small animals (jPET-RD), we performed Monte Carlo simulations using the GATE based on Geant4 and investigated the influence of the proximity of the detectors to the target object on the sensitivity and count rate performance. The jPET-RD is based on a large-size depth-of-interaction (DOI) block detector that consists of a 4-layered array of 32 x 32 LSO crystals (1.4 mm x 1.4 mm x 4.5 mm) and a 256-ch flat panel position-sensitive photomultiplier tube. In this work, three detector geometries were simulated: two rings of six detector blocks arranged in a hexagonal pattern (FOV 85 mm in diameter) and four detector blocks arranged in a tetragonal pattern (FOV 49 mm in diameter) and in an overlapped tetragonal pattern (FOV 38 mm in diameter). The simulation results showed that the smaller bore geometry can provide higher sensitivity because of its larger solid angle. Although it clearly affected the noise equivalent count rate (NECR) due to its high dead-time, parallel readout with appropriate anode segmentation improved the NECR at 20 MBq by a factor of 1.1 to 1.4 compared with the case of 256-ch anodes read out by one front-end circuit. (author)

  11. Geiger-Mueller haloid counter dead time dependence on counting rate

    International Nuclear Information System (INIS)

    Onishchenko, A.M.; Tsvetkov, A.A.

    1980-01-01

    The experimental dependences of the dead time of Geiger counters (SBM-19, SBM-20, SBM-21 and SGM-19) on the loading, are presented. The method of two sources has been used to determine the dead time counters of increased stability. The counters are switched on according to the usually used circuit of discrete counting with loading resistance of 50 MOhm and the separating capacity of 10 pF. Voltage pulses are given to the counting device with the time of resolution of 100 ns, discrimenation threshold 3 V, input resistance 3.6 Ω and the input capacity-15 pF. The time constant of the counter RC-circuit is 50 μs

  12. Optimization of the ATLAS (s)MDT readout electronics for high counting rates

    Energy Technology Data Exchange (ETDEWEB)

    Kortner, Oliver; Kroha, Hubert; Nowak, Sebastian; Schmidt-Sommerfeld, Korbinian [Max-Planck-Institut fuer Physik (Werner-Heisenberg-Institut), Foehringer Ring 6, 80805 Muenchen (Germany)

    2016-07-01

    In the ATLAS muon spectrometer, Monitored Drift Tube (MDT) chambers are used for precise muon track measurement. For the high background rates expected at HL-LHC, which are mainly due to neutrons and photons produced by interactions of the proton collision products in the detector and shielding, new small-diameter muon drift tube (sMDT)-chambers with half the drift tube diameter of the MDT-chambers and ten times higher rate capability have been developed. The standard MDT readout electronics uses bipolar shaping in front of a discriminator. This shaping leads to an undershoot of same charge but opposite polarity following each pulse. With count rates also the probability of having the subsequent pulse in this undershoot increases, which leads to losses in efficiency and spatial resolution. In order to decrease this effect, discrete prototype electronics including Baseline Restoration has been developed. Results of their tests and data taken with them during muon beamtime measurements at CERN's Gamma Irradiation Facility will be presented. which causes a deterioration of signal pulses by preceding background hits, leading to losses in muon efficiency and drift tube spatial resolution. In order to mitigate these so-called signal pile-up effects, new readout electronics with active baseline restoration (BLR) is under development. Discrete prototype electronics with BLR functionality has been tested in laboratory measurements and in the Gamma Irradiation Facility at CERN under high γ-irradiation rates. Results of the measurements are presented.

  13. 30 CFR 75.601-3 - Short circuit protection; dual element fuses; current ratings; maximum values.

    Science.gov (United States)

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Short circuit protection; dual element fuses... Trailing Cables § 75.601-3 Short circuit protection; dual element fuses; current ratings; maximum values. Dual element fuses having adequate current-interrupting capacity shall meet the requirements for short...

  14. Detectors for proton counting. Si-APD and scintillation detectors

    International Nuclear Information System (INIS)

    Kishimoto, Shunji

    2008-01-01

    Increased intensity of synchrotron radiation requests users to prepare photon pulse detectors having higher counting rates. As detectors for photon counting, silicon-avalanche photodiode (Si-APD) and scintillation detectors were chosen for the fifth series of detectors. Principle of photon detection by pulse and need of amplification function of the detector were described. Structure and working principle, high counting rate measurement system, bunch of electrons vs. counting rate, application example of NMR time spectroscopy measurement and comments for users were described for the Si-APD detector. Structure of scintillator and photomultiplier tube, characteristics of scintillator and performance of detector were shown for the NaI detector. Future development of photon pulse detectors was discussed. (T. Tanaka)

  15. 15Mcps photon-counting X-ray computed tomography system using a ZnO-MPPC detector and its application to gadolinium imaging.

    Science.gov (United States)

    Sato, Eiichi; Sugimura, Shigeaki; Endo, Haruyuki; Oda, Yasuyuki; Abudurexiti, Abulajiang; Hagiwara, Osahiko; Osawa, Akihiro; Matsukiyo, Hiroshi; Enomoto, Toshiyuki; Watanabe, Manabu; Kusachi, Shinya; Sato, Shigehiro; Ogawa, Akira; Onagawa, Jun

    2012-01-01

    15Mcps photon-counting X-ray computed tomography (CT) system is a first-generation type and consists of an X-ray generator, a turntable, a translation stage, a two-stage controller, a detector consisting of a 2mm-thick zinc-oxide (ZnO) single-crystal scintillator and an MPPC (multipixel photon counter) module, a counter card (CC), and a personal computer (PC). High-speed photon counting was carried out using the detector in the X-ray CT system. The maximum count rate was 15Mcps (mega counts per second) at a tube voltage of 100kV and a tube current of 1.95mA. Tomography is accomplished by repeated translations and rotations of an object, and projection curves of the object are obtained by the translation. The pulses of the event signal from the module are counted by the CC in conjunction with the PC. The minimum exposure time for obtaining a tomogram was 15min, and photon-counting CT was accomplished using gadolinium-based contrast media. Copyright © 2011 Elsevier Ltd. All rights reserved.

  16. More accurate thermal neutron coincidence counting technique

    International Nuclear Information System (INIS)

    Baron, N.

    1978-01-01

    Using passive thermal neutron coincidence counting techniques, the accuracy of nondestructive assays of fertile material can be improved significantly using a two-ring detector. It was shown how the use of a function of the coincidence count rate ring-ratio can provide a detector response rate that is independent of variations in neutron detection efficiency caused by varying sample moderation. Furthermore, the correction for multiplication caused by SF- and (α,n)-neutrons is shown to be separable into the product of a function of the effective mass of 240 Pu (plutonium correction) and a function of the (α,n) reaction probability (matrix correction). The matrix correction is described by a function of the singles count rate ring-ratio. This correction factor is empirically observed to be identical for any combination of PuO 2 powder and matrix materials SiO 2 and MgO because of the similar relation of the (α,n)-Q value and (α,n)-reaction cross section among these matrix nuclei. However the matrix correction expression is expected to be different for matrix materials such as Na, Al, and/or Li. Nevertheless, it should be recognized that for comparison measurements among samples of similar matrix content, it is expected that some function of the singles count rate ring-ratio can be defined to account for variations in the matrix correction due to differences in the intimacy of mixture among the samples. Furthermore the magnitude of this singles count rate ring-ratio serves to identify the contaminant generating the (α,n)-neutrons. Such information is useful in process control

  17. Unintentional carbon monoxide poisoning hospitalization and emergency department counts and rates by county, year, and fire-relatedness among California residents,2000-2007

    Data.gov (United States)

    California Environmental Health Tracking Program — This dataset contains case counts, rates, and confidence intervals of unintentional carbon monoxide poisoning (CO) inpatient hospitalizations and emergency...

  18. The absolute counting of red cell-derived microparticles with red cell bead by flow rate based assay.

    Science.gov (United States)

    Nantakomol, Duangdao; Imwong, Malika; Soontarawirat, Ingfar; Kotjanya, Duangporn; Khakhai, Chulalak; Ohashi, Jun; Nuchnoi, Pornlada

    2009-05-01

    Activation of red blood cell is associated with the formation of red cell-derived microparticles (RMPs). Analysis of circulating RMPs is becoming more refined and clinically useful. A quantitative Trucount tube method is the conventional method uses for quantitating RMPs. In this study, we validated a quantitative method called "flow rate based assay using red cell bead (FCB)" to measure circulating RMPs in the peripheral blood of healthy subjects. Citrated blood samples collected from 30 cases of healthy subjects were determined the RMPs count by using double labeling of annexin V-FITC and anti-glycophorin A-PE. The absolute RMPs numbers were measured by FCB, and the results were compared with the Trucount or with flow rate based calibration (FR). Statistical correlation and agreement were analyzed using linear regression and Bland-Altman analysis. There was no significant difference in the absolute number of RMPs quantitated by FCB when compared with those two reference methods including the Trucount tube and FR method. The absolute RMPs count obtained from FCB method was highly correlated with those obtained from Trucount tube (r(2) = 0.98, mean bias 4 cell/microl, limit of agreement [LOA] -20.3 to 28.3 cell/microl), and FR method (r(2) = 1, mean bias 10.3 cell/microl, and LOA -5.5 to 26.2 cell/microl). This study demonstrates that FCB is suitable and more affordable for RMPs quantitation in the clinical samples. This method is a low cost and interchangeable to latex bead-based method for generating the absolute counts in the resource-limited areas. (c) 2008 Clinical Cytometry Society.

  19. Improving the counting efficiency in time-correlated single photon counting experiments by dead-time optimization

    Energy Technology Data Exchange (ETDEWEB)

    Peronio, P.; Acconcia, G.; Rech, I.; Ghioni, M. [Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milano (Italy)

    2015-11-15

    Time-Correlated Single Photon Counting (TCSPC) has been long recognized as the most sensitive method for fluorescence lifetime measurements, but often requiring “long” data acquisition times. This drawback is related to the limited counting capability of the TCSPC technique, due to pile-up and counting loss effects. In recent years, multi-module TCSPC systems have been introduced to overcome this issue. Splitting the light into several detectors connected to independent TCSPC modules proportionally increases the counting capability. Of course, multi-module operation also increases the system cost and can cause space and power supply problems. In this paper, we propose an alternative approach based on a new detector and processing electronics designed to reduce the overall system dead time, thus enabling efficient photon collection at high excitation rate. We present a fast active quenching circuit for single-photon avalanche diodes which features a minimum dead time of 12.4 ns. We also introduce a new Time-to-Amplitude Converter (TAC) able to attain extra-short dead time thanks to the combination of a scalable array of monolithically integrated TACs and a sequential router. The fast TAC (F-TAC) makes it possible to operate the system towards the upper limit of detector count rate capability (∼80 Mcps) with reduced pile-up losses, addressing one of the historic criticisms of TCSPC. Preliminary measurements on the F-TAC are presented and discussed.

  20. Aperture and counting rate of rectangular telescopes for single and multiple parallel particles. [Spark chamber telescopes

    Energy Technology Data Exchange (ETDEWEB)

    D' Ettorre Piazzoli, B; Mannocchi, G [Consiglio Nazionale delle Ricerche, Turin (Italy). Lab. di Cosmo-Geofisica; Melone, S [Istituto di Fisica dell' Universita, Ancona, Italy; Picchi, P; Visentin, R [Comitato Nazionale per l' Energia Nucleare, Frascati (Italy). Laboratori Nazionali di Frascati

    1976-06-01

    Expressions for the counting rate of rectangular telescopes in the case of single as well as multiple particles are given. The aperture for single particles is obtained in the form of a double integral and analytical solutions are given for some cases. The intensity for different multiplicities of parallel particles is related to the geometry of the detectors and to the features of the radiation. This allows an absolute comparison between the data recorded by different devices.

  1. ReplacementMatrix: a web server for maximum-likelihood estimation of amino acid replacement rate matrices.

    Science.gov (United States)

    Dang, Cuong Cao; Lefort, Vincent; Le, Vinh Sy; Le, Quang Si; Gascuel, Olivier

    2011-10-01

    Amino acid replacement rate matrices are an essential basis of protein studies (e.g. in phylogenetics and alignment). A number of general purpose matrices have been proposed (e.g. JTT, WAG, LG) since the seminal work of Margaret Dayhoff and co-workers. However, it has been shown that matrices specific to certain protein groups (e.g. mitochondrial) or life domains (e.g. viruses) differ significantly from general average matrices, and thus perform better when applied to the data to which they are dedicated. This Web server implements the maximum-likelihood estimation procedure that was used to estimate LG, and provides a number of tools and facilities. Users upload a set of multiple protein alignments from their domain of interest and receive the resulting matrix by email, along with statistics and comparisons with other matrices. A non-parametric bootstrap is performed optionally to assess the variability of replacement rate estimates. Maximum-likelihood trees, inferred using the estimated rate matrix, are also computed optionally for each input alignment. Finely tuned procedures and up-to-date ML software (PhyML 3.0, XRATE) are combined to perform all these heavy calculations on our clusters. http://www.atgc-montpellier.fr/ReplacementMatrix/ olivier.gascuel@lirmm.fr Supplementary data are available at http://www.atgc-montpellier.fr/ReplacementMatrix/

  2. Principles of correlation counting

    International Nuclear Information System (INIS)

    Mueller, J.W.

    1975-01-01

    A review is given of the various applications which have been made of correlation techniques in the field of nuclear physics, in particular for absolute counting. Whereas in most cases the usual coincidence method will be preferable for its simplicity, correlation counting may be the only possible approach in such cases where the two radiations of the cascade cannot be well separated or when there is a longliving intermediate state. The measurement of half-lives and of count rates of spurious pulses is also briefly discussed. The various experimental situations lead to different ways the correlation method is best applied (covariance technique with one or with two detectors, application of correlation functions, etc.). Formulae are given for some simple model cases, neglecting dead-time corrections

  3. Monitoring Milk Somatic Cell Counts

    Directory of Open Access Journals (Sweden)

    Gheorghe Şteţca

    2014-11-01

    Full Text Available The presence of somatic cells in milk is a widely disputed issue in milk production sector. The somatic cell counts in raw milk are a marker for the specific cow diseases such as mastitis or swollen udder. The high level of somatic cells causes physical and chemical changes to milk composition and nutritional value, and as well to milk products. Also, the mastitic milk is not proper for human consumption due to its contribution to spreading of certain diseases and food poisoning. According to these effects, EU Regulations established the maximum threshold of admitted somatic cells in raw milk to 400000 cells / mL starting with 2014. The purpose of this study was carried out in order to examine the raw milk samples provided from small farms, industrial type farms and milk processing units. There are several ways to count somatic cells in milk but the reference accepted method is the microscopic method described by the SR EN ISO 13366-1/2008. Generally samples registered values in accordance with the admissible limit. By periodical monitoring of the somatic cell count, certain technological process issues are being avoided and consumer’s health ensured.

  4. Bayesian approach in MN low dose of radiation counting

    International Nuclear Information System (INIS)

    Serna Berna, A.; Alcaraz, M.; Acevedo, C.; Navarro, J. L.; Alcanzar, M. D.; Canteras, M.

    2006-01-01

    The Micronucleus assay in lymphocytes is a well established technique for the assessment of genetic damage induced by ionizing radiation. Due to the presence of a natural background of MN the net MN is obtained by subtracting this value to the gross value. When very low doses of radiation are given the induced MN is close even lower than the predetermined background value. Furthermore, the damage distribution induced by the radiation follows a Poisson probability distribution. These two facts pose a difficult task to obtain the net counting rate in the exposed situations. It is possible to overcome this problem using a bayesian approach, in which the selection of a priori distributions for the background and net counting rate plays an important role. In the present work we make a detailed analysed using bayesian theory to infer the net counting rate in two different situations: a) when the background is known for an individual sample, using exact value value for the background and Jeffreys prior for the net counting rate, and b) when the background is not known and we make use of a population background distribution as background prior function and constant prior for the net counting rate. (Author)

  5. Front-end counting mode electronics for CdZnTe sensor readout

    CERN Document Server

    Moraes, Danielle; Kaplon, Jan

    2004-01-01

    The development of a front-end circuit optimized for CdZnTe detector readout, implemented in 0.25 mu m CMOS technology, is reported. The ASIC comprises 17 channels of a charge sensitive amplifier with an active feedback, followed by a gain-shaper stage and a discriminator with a 5 bit fine-tune DAC. The signal from the discriminator is sensed by a 25 ns mono-stable circuit and an 18-bit static ripple- counter. The channel architecture is optimized for the detector characteristics in order to achieve the best energy resolution at a maximum counting rate of 2 million counts/second. The amplifier shows a linear sensitivity of 24 mV/fC with 50 ns peaking time and an equivalent noise charge of about 650 e/sup -/, for a detector capacitance of 10 pF. When connected to a 3*3*7 mm/sup 3/ CdZnTe detector the amplifier gain is about 8 mV/keV with a noise around 3.6 keV.

  6. Loss-Free Counting with Digital Signal Processors

    International Nuclear Information System (INIS)

    Markku Koskelo; Dave Hall; Martin Moslinger

    2000-01-01

    Loss-free-counting (LFC) techniques have frequently been used with traditional analog pulse processing systems to compensate for the time or pulses lost when a spectroscopy system is unavailable (busy) for processing an accepted pulse. With the availability of second-generation digital signal processing (DSP) electronics that offer a significantly improved performance for both high and low count rate applications, the LFC technique has been revisited. Specific attention was given to the high and ultra-high count rate behavior, using high-purity germanium (HPGe) detectors with both transistor reset preamplifiers (TRP) and conventional RC preamplifiers. The experiments conducted for this work show that the known LFC techniques further benefit when combined with modern DSP pulse shaping

  7. Maximum heart rate in brown trout (Salmo trutta fario) is not limited by firing rate of pacemaker cells.

    Science.gov (United States)

    Haverinen, Jaakko; Abramochkin, Denis V; Kamkin, Andre; Vornanen, Matti

    2017-02-01

    Temperature-induced changes in cardiac output (Q̇) in fish are largely dependent on thermal modulation of heart rate (f H ), and at high temperatures Q̇ collapses due to heat-dependent depression of f H This study tests the hypothesis that firing rate of sinoatrial pacemaker cells sets the upper thermal limit of f H in vivo. To this end, temperature dependence of action potential (AP) frequency of enzymatically isolated pacemaker cells (pacemaker rate, f PM ), spontaneous beating rate of isolated sinoatrial preparations (f SA ), and in vivo f H of the cold-acclimated (4°C) brown trout (Salmo trutta fario) were compared under acute thermal challenges. With rising temperature, f PM steadily increased because of the acceleration of diastolic depolarization and shortening of AP duration up to the break point temperature (T BP ) of 24.0 ± 0.37°C, at which point the electrical activity abruptly ceased. The maximum f PM at T BP was much higher [193 ± 21.0 beats per minute (bpm)] than the peak f SA (94.3 ± 6.0 bpm at 24.1°C) or peak f H (76.7 ± 2.4 at 15.7 ± 0.82°C) (P brown trout in vivo. Copyright © 2017 the American Physiological Society.

  8. High count-rate study of two TES x-ray microcalorimeters with different transition temperatures

    Science.gov (United States)

    Lee, Sang-Jun; Adams, Joseph S.; Bandler, Simon R.; Betancourt-Martinez, Gabriele L.; Chervenak, James A.; Eckart, Megan E.; Finkbeiner, Fred M.; Kelley, Richard L.; Kilbourne, Caroline A.; Porter, Frederick S.; Sadleir, John E.; Smith, Stephen J.; Wassell, Edward J.

    2017-10-01

    We have developed transition-edge sensor (TES) microcalorimeter arrays with high count-rate capability and high energy resolution to carry out x-ray imaging spectroscopy observations of various astronomical sources and the Sun. We have studied the dependence of the energy resolution and throughput (fraction of processed pulses) on the count rate for such microcalorimeters with two different transition temperatures (T c). Devices with both transition temperatures were fabricated within a single microcalorimeter array directly on top of a solid substrate where the thermal conductance of the microcalorimeter is dependent upon the thermal boundary resistance between the TES sensor and the dielectric substrate beneath. Because the thermal boundary resistance is highly temperature dependent, the two types of device with different T cs had very different thermal decay times, approximately one order of magnitude different. In our earlier report, we achieved energy resolutions of 1.6 and 2.3 eV at 6 keV from lower and higher T c devices, respectively, using a standard analysis method based on optimal filtering in the low flux limit. We have now measured the same devices at elevated x-ray fluxes ranging from 50 Hz to 1000 Hz per pixel. In the high flux limit, however, the standard optimal filtering scheme nearly breaks down because of x-ray pile-up. To achieve the highest possible energy resolution for a fixed throughput, we have developed an analysis scheme based on the so-called event grade method. Using the new analysis scheme, we achieved 5.0 eV FWHM with 96% throughput for 6 keV x-rays of 1025 Hz per pixel with the higher T c (faster) device, and 5.8 eV FWHM with 97% throughput with the lower T c (slower) device at 722 Hz.

  9. Outpacing the Anthropocene: New Constraints for the Rate of Carbon Release at the Paleocene-Eocene Thermal Maximum

    Science.gov (United States)

    Wright, J. D.; Schaller, M. F.

    2012-12-01

    The Paleocene/Eocene Thermal Maximum (PETM) Carbon Isotope Excursion (CIE) is linked to benthic foraminiferal extinction and excursion taxa in planktonic foraminifera and calcareous nannofossils. Previous studies have used integrated bio-magneto-stratigraphies, cycle counting, and extraterrestrial 3He accumulation rates to produce a range of estimates for the duration of the initial onset of the PETM CIE between 750 years to 30 kyr. Durations for the total release time (onset to initiation of recovery) range from 45 to 95 kyr. Uncertainty in the timing of the onset of the PETM CIE prevents the identification of a causal mechanism, and hence understanding the biological responses. Recent work on the Paleocene/Eocene Marlboro Clay has unveiled the presence of regular couplets (~2 cm) expressed in multiple cores and exposures throughout the Atlantic Coastal Plain. Specifically, the Millville and newly recovered Wilson Lake B cores contain 750 and 660 layers through the CIE, respectively. These couplets have corresponding oxygen stable isotope cycles, arguing for a climatic origin. Orbital and millennial periodicities are far too long to explain the ~750 layers identified in the Millville core. Seasonal insolation is the only regular climate cycle that can plausibly account for the observed δ18O amplitudes (~1‰, with some cycles up to 2‰) and layer counts. Seasonal freshwater input can also augment the cyclic oscillations in δ18O, but the majority of the variability is most plausibly ascribed to temperature. Wilson Lake B and Millville have total δ13C excursions of -5 and -4.5‰ respectively, as well as highly expanded sections of the PETM CIE. In the Millville core, high-resolution, bulk stable isotope records show a 3.5‰ δ13C decrease over 12 layers across the PETM CIE onset. Concomitant with this δ13C decrease is a sharp drop in CaCO3. Decreases in both proxies require a large, sudden release of isotopically light carbon. The couplet chronology indicates

  10. Control of the neutron detector count rate by optical imaging

    International Nuclear Information System (INIS)

    Roquemore, A.L.; Johnson, L.C.

    1992-01-01

    The signal processing electronics used for the NE451 detectors on the TFTR multichannel neutron collimator are presently showing saturation effects at high counting rates equivalent to neutron yields of ∼10 16 n/s. While nonlinearity due to pulse pileup can be corrected for in most present TFTR experiments, additional steps are required for neutron source strengths above ∼3x10 16 n/s. These pulse pileup effects could be reduced by inserting sleeves in the collimator shielding to reduce the neutron flux in the vicinity of the detectors or by reducing the volume of detector exposed to the flux. We describe a novel method of avoiding saturation by optically controlling the number neutron events processed by the detector electronics. Because of the optical opacity of the ZnS-plastic detectors such as NE451, photons from a proton-recoil scintillation arise from a spatially localized area of the detector. By imaging a selected portion of the detector onto a photomultiplier, we reduce the effective volume of the detector in a controllable, reversible way. A prototype system, consisting of a focusing lens, a field lens, and a variable aperture, has been constructed. Results of laboratory feasibility tests are presented

  11. Standardization of Ga-68 by coincidence measurements, liquid scintillation counting and 4πγ counting.

    Science.gov (United States)

    Roteta, Miguel; Peyres, Virginia; Rodríguez Barquero, Leonor; García-Toraño, Eduardo; Arenillas, Pablo; Balpardo, Christian; Rodrígues, Darío; Llovera, Roberto

    2012-09-01

    The radionuclide (68)Ga is one of the few positron emitters that can be prepared in-house without the use of a cyclotron. It disintegrates to the ground state of (68)Zn partially by positron emission (89.1%) with a maximum energy of 1899.1 keV, and partially by electron capture (10.9%). This nuclide has been standardized in the frame of a cooperation project between the Radionuclide Metrology laboratories from CIEMAT (Spain) and CNEA (Argentina). Measurements involved several techniques: 4πβ-γ coincidences, integral gamma counting and Liquid Scintillation Counting using the triple to double coincidence ratio and the CIEMAT/NIST methods. Given the short half-life of the radionuclide assayed, a direct comparison between results from both laboratories was excluded and a comparison of experimental efficiencies of similar NaI detectors was used instead. Copyright © 2012 Elsevier Ltd. All rights reserved.

  12. HEPS-BPIX, a single photon counting pixel detector with a high frame rate for the HEPS project

    Energy Technology Data Exchange (ETDEWEB)

    Wei, Wei, E-mail: weiw@ihep.ac.cn [Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049 (China); State Key Laboratory of Particle Detection and Electronics, Beijing 100049 (China); Zhang, Jie; Ning, Zhe; Lu, Yunpeng; Fan, Lei; Li, Huaishen; Jiang, Xiaoshan; Lan, Allan K.; Ouyang, Qun; Wang, Zheng; Zhu, Kejun; Chen, Yuanbo [Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049 (China); State Key Laboratory of Particle Detection and Electronics, Beijing 100049 (China); Liu, Peng [Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049 (China)

    2016-11-01

    China's next generation light source, named the High Energy Photon Source (HEPS), is currently under construction. HEPS-BPIX (HEPS-Beijing PIXel) is a dedicated pixel readout chip that operates in single photon counting mode for X-ray applications in HEPS. Designed using CMOS 0.13 µm technology, the chip contains a matrix of 104×72 pixels. Each pixel measures 150 µm×150 µm and has a counting depth of 20 bits. A bump-bonded prototyping detector module with a 300-µm thick silicon sensor was tested in the beamline of Beijing Synchrotron Radiation Facility. A fast stream of X-ray images was demonstrated, and a frame rate of 1.2 kHz was proven, with a negligible dead time. The test results showed an equivalent noise charge of 115 e{sup −} rms after bump bonding and a threshold dispersion of 55 e{sup −} rms after calibration.

  13. Computed neutron coincidence counting applied to passive waste assay

    Energy Technology Data Exchange (ETDEWEB)

    Bruggeman, M.; Baeten, P.; De Boeck, W.; Carchon, R. [Nuclear Research Centre, Mol (Belgium)

    1997-11-01

    Neutron coincidence counting applied for the passive assay of fissile material is generally realised with dedicated electronic circuits. This paper presents a software based neutron coincidence counting method with data acquisition via a commercial PC-based Time Interval Analyser (TIA). The TIA is used to measure and record all time intervals between successive pulses in the pulse train up to count-rates of 2 Mpulses/s. Software modules are then used to compute the coincidence count-rates and multiplicity related data. This computed neutron coincidence counting (CNCC) offers full access to all the time information contained in the pulse train. This paper will mainly concentrate on the application and advantages of CNCC for the non-destructive assay of waste. An advanced multiplicity selective Rossi-alpha method is presented and its implementation via CNCC demonstrated. 13 refs., 4 figs., 2 tabs.

  14. Computed neutron coincidence counting applied to passive waste assay

    International Nuclear Information System (INIS)

    Bruggeman, M.; Baeten, P.; De Boeck, W.; Carchon, R.

    1997-01-01

    Neutron coincidence counting applied for the passive assay of fissile material is generally realised with dedicated electronic circuits. This paper presents a software based neutron coincidence counting method with data acquisition via a commercial PC-based Time Interval Analyser (TIA). The TIA is used to measure and record all time intervals between successive pulses in the pulse train up to count-rates of 2 Mpulses/s. Software modules are then used to compute the coincidence count-rates and multiplicity related data. This computed neutron coincidence counting (CNCC) offers full access to all the time information contained in the pulse train. This paper will mainly concentrate on the application and advantages of CNCC for the non-destructive assay of waste. An advanced multiplicity selective Rossi-alpha method is presented and its implementation via CNCC demonstrated. 13 refs., 4 figs., 2 tabs

  15. Maximum-likelihood fitting of data dominated by Poisson statistical uncertainties

    International Nuclear Information System (INIS)

    Stoneking, M.R.; Den Hartog, D.J.

    1996-06-01

    The fitting of data by χ 2 -minimization is valid only when the uncertainties in the data are normally distributed. When analyzing spectroscopic or particle counting data at very low signal level (e.g., a Thomson scattering diagnostic), the uncertainties are distributed with a Poisson distribution. The authors have developed a maximum-likelihood method for fitting data that correctly treats the Poisson statistical character of the uncertainties. This method maximizes the total probability that the observed data are drawn from the assumed fit function using the Poisson probability function to determine the probability for each data point. The algorithm also returns uncertainty estimates for the fit parameters. They compare this method with a χ 2 -minimization routine applied to both simulated and real data. Differences in the returned fits are greater at low signal level (less than ∼20 counts per measurement). the maximum-likelihood method is found to be more accurate and robust, returning a narrower distribution of values for the fit parameters with fewer outliers

  16. Measurement of binding of ascorbic acid to myrosinase by rate of dialysis

    International Nuclear Information System (INIS)

    Ohtsuru, Masaru; Hata, Tadao

    1975-01-01

    The activation mechanism of myrosinase by L-ascorbic acid depends on the slight conformational change of enzyme protein induced by ascorbic acid. Ascorbic acid binds to enzyme like Michaelis-complex, and then the value of Km had been evaluated to be 1 x 10 -3 M. The authors determined the binding constant and the number of binding sites using dialysis rate technique. Rate dialysis was carried out with a dialysis cell, and the ordinary cellophane tubing membrane was used. ( 14 C) - ascorbic acid was added, and counted by liquid scintillation counting. By the time course of two dialysis rate measurement with and without enzyme. The concentrations of free and bound ascorbic acids were counted. From the results, the enzyme was activated to the maximum level at 10 -3 M of ascorbic acid, and four molecules of ascorbic acid bound to the enzyme on Kd=0.1x10 -4 M. However, when more than 4 molecules of L-ascorbic acid bound to the enzyme, Kd increased to 0.9x10 -4 M, and L-ascorbic acid acted as an inhibitor. (Kubatake, H.)

  17. Clinical evaluation of a simple uroflowmeter for categorization of maximum urinary flow rate

    Directory of Open Access Journals (Sweden)

    Simon Pridgeon

    2007-01-01

    Full Text Available Objective: To evaluate the accuracy and diagnostic usefulness of a disposable flowmeter consisting of a plastic funnel with a spout divided into three chambers. Materials and Methods: Men with lower urinary tract symptoms (LUTS voided sequentially into a standard flowmeter and the funnel device recording maximum flow rate (Q max and voided volume (V void . The device was precalibrated such that filling of the bottom, middle and top chambers categorized maximum input flows as 15 ml s -1 respectively. Subjects who agreed to use the funnel device at home obtained readings of flow category and V void twice daily for seven days. Results: A single office reading in 46 men using the device showed good agreement with standard measurement of Q max for V void > 150 ml (Kappa = 0.68. All 14 men whose void reached the top chamber had standard Q max > 15 ml s -1 (PPV = 100%, NPV = 72% whilst eight of 12 men whose void remained in the bottom chamber had standard Q max < 10 ml s -1 (PPV = 70%, NPV = 94%. During multiple home use by 14 men the device showed moderate repeatability (Kappa = 0.58 and correctly categorized Q max in comparison to standard measurement for 12 (87% men. Conclusions: This study suggests that the device has sufficient accuracy and reliability for initial flow rate assessment in men with LUTS. The device can provide a single measurement or alternatively multiple home measurements to categorize men with Q max < 15 ml s -1 .

  18. Aspartic acid racemization rate in narwhal (Monodon monoceros) eye lens nuclei estimated by counting of growth layers in tusks

    DEFF Research Database (Denmark)

    Garde, Eva; Heide-Jørgensen, Mads Peter; Ditlevsen, Susanne

    2012-01-01

    Ages of marine mammals have traditionally been estimated by counting dentinal growth layers in teeth. However, this method is difficult to use on narwhals (Monodon monoceros) because of their special tooth structures. Alternative methods are therefore needed. The aspartic acid racemization (AAR......) technique has been used in age estimation studies of cetaceans, including narwhals. The purpose of this study was to estimate a species-specific racemization rate for narwhals by regressing aspartic acid D/L ratios in eye lens nuclei against growth layer groups in tusks (n=9). Two racemization rates were...

  19. Experimental verification of preset time count rate meters based on adaptive digital signal processing algorithms

    Directory of Open Access Journals (Sweden)

    Žigić Aleksandar D.

    2005-01-01

    Full Text Available Experimental verifications of two optimized adaptive digital signal processing algorithms implemented in two pre set time count rate meters were per formed ac cording to appropriate standards. The random pulse generator realized using a personal computer, was used as an artificial radiation source for preliminary system tests and performance evaluations of the pro posed algorithms. Then measurement results for background radiation levels were obtained. Finally, measurements with a natural radiation source radioisotope 90Sr-90Y, were carried out. Measurement results, con ducted without and with radio isotopes for the specified errors of 10% and 5% showed to agree well with theoretical predictions.

  20. Liquid scintillation counting of chlorophyll

    International Nuclear Information System (INIS)

    Fric, F.; Horickova, B.; Haspel-Horvatovic, E.

    1975-01-01

    A precise and reproducible method of liquid scintillation counting was worked out for measuring the radioactivity of 14 C-labelled chlorophyll a and chlorophyll b solutions without previous bleaching. The spurious count rate caused by luminescence of the scintillant-chlorophyll system is eliminated by using a suitable scintillant and by measuring the radioactivity at 4 to 8 0 C after an appropriate time of dark adaptation. Bleaching of the chlorophyll solutions is necessary only for measuring of very low radioactivity. (author)

  1. Development of a high-count-rate neutron detector with position sensitivity and high efficiency

    International Nuclear Information System (INIS)

    Nelson, R.; Sandoval, J.

    1996-01-01

    While the neutron scattering community is bombarded with hints of new technologies that may deliver detectors with high-count-rate capability, high efficiency, gamma-ray insensitivity, and high resolution across large areas, only the time-tested, gas-filled 3 He and scintillation detectors are in widespread use. Future spallation sources with higher fluxes simply must exploit some of the advanced detector schemes that are as yet unproved as production systems. Technologies indicating promise as neutron detectors include pixel arrays of amorphous silicon, silicon microstrips, microstrips with gas, and new scintillation materials. This project sought to study the competing neutron detector technologies and determine which or what combination will lead to a production detector system well suited for use at a high-intensity neutron scattering source

  2. Spectrometry with high count rate for the study of the soft X-rays. Application for the plasma of WEGA

    International Nuclear Information System (INIS)

    Brouquet, P.

    1979-04-01

    The plasma of the WEGA torus, whose electron temperature varies between 0.5 and 1 keV, emits electromagnetic radiation extending to wavelengths of the order of 1A. Different improvements performed on a semi-conductor spectrometer have permitted the study of this emission in the soft X ray region (1 keV - 30 keV) at a count rate of 3.10 5 counts/s with an energy resolution of 350 eV. For each plasma shot, this diagnostic gives 4 measurements of the plasma electron temperature and of the effective charge, Zeff, with a time resolution of 5 ms. The values of the electron temperature and of the effective charge derived from the study of soft X rays are in agreement with those given by other diagnostic methods [fr

  3. Characterization of photon-counting multislit breast tomosynthesis.

    Science.gov (United States)

    Berggren, Karl; Cederström, Björn; Lundqvist, Mats; Fredenberg, Erik

    2018-02-01

    It has been shown that breast tomosynthesis may improve sensitivity and specificity compared to two-dimensional mammography, resulting in increased detection-rate of cancers or lowered call-back rates. The purpose of this study is to characterize a spectral photon-counting multislit breast tomosynthesis system that is able to do single-scan spectral imaging with multiple collimated x-ray beams. The system differs in many aspects compared to conventional tomosynthesis using energy-integrating flat-panel detectors. The investigated system was a prototype consisting of a dual-threshold photon-counting detector with 21 collimated line detectors scanning across the compressed breast. A review of the system is done in terms of detector, acquisition geometry, and reconstruction methods. Three reconstruction methods were used, simple back-projection, filtered back-projection and an iterative algebraic reconstruction technique. The image quality was evaluated by measuring the modulation transfer-function (MTF), normalized noise-power spectrum, detective quantum-efficiency (DQE), and artifact spread-function (ASF) on reconstructed spectral tomosynthesis images for a total-energy bin (defined by a low-energy threshold calibrated to remove electronic noise) and for a high-energy bin (with a threshold calibrated to split the spectrum in roughly equal parts). Acquisition was performed using a 29 kVp W/Al x-ray spectrum at a 0.24 mGy exposure. The difference in MTF between the two energy bins was negligible, that is, there was no energy dependence on resolution. The MTF dropped to 50% at 1.5 lp/mm to 2.3 lp/mm in the scan direction and 2.4 lp/mm to 3.3 lp/mm in the slit direction, depending on the reconstruction method. The full width at half maximum of the ASF was found to range from 13.8 mm to 18.0 mm for the different reconstruction methods. The zero-frequency DQE of the system was found to be 0.72. The fraction of counts in the high-energy bin was measured to be 59% of the

  4. Counting statistics and loss corrections for the APS

    International Nuclear Information System (INIS)

    Lee, W.K.; Mills, D.M.

    1992-01-01

    It has been suggested that for timing experiments, it might be advantageous to arrange the bunches in the storage ring in an asymmetrical mode. In this paper, we determine the counting losses from pulsed x-ray sources from basic probabilistic arguments and from Poisson statistics. In particular the impact on single photon counting losses of a variety of possible filling modes for the Advanced Photon Source (APS) is examined. For bunches of equal current, a loss of 10% occurs whenever the count rate exceeds 21% of the bunch repetition rate. This changes slightly when bunches containing unequal numbers of particles are considered. The results are applied to several common detector/electronics systems

  5. Counting statistics and loss corrections for the APS

    International Nuclear Information System (INIS)

    Lee, W.K.; Mills, D.M.

    1992-01-01

    It has been suggested that for timing experiments, it might be advantageous to arrange the bunches in the storage ring in an asymmetrical mode. In this paper, we determine the counting losses from pulsed x-ray sources from basic probabilistic arguments and from Poisson statistics. In particular the impact on single-photon counting losses of a variety of possible filling modes for the Advanced Photon Source (APS) is examined. For bunches of equal current, a loss of 10% occurs whenever the count rate exceeds 21% of the bunch repetition rate. This changes slightly when bunches containing unequal numbers of particles are considered. The results are applied to several common detector/electronics systems

  6. Photon Counting Using Edge-Detection Algorithm

    Science.gov (United States)

    Gin, Jonathan W.; Nguyen, Danh H.; Farr, William H.

    2010-01-01

    New applications such as high-datarate, photon-starved, free-space optical communications require photon counting at flux rates into gigaphoton-per-second regimes coupled with subnanosecond timing accuracy. Current single-photon detectors that are capable of handling such operating conditions are designed in an array format and produce output pulses that span multiple sample times. In order to discern one pulse from another and not to overcount the number of incoming photons, a detection algorithm must be applied to the sampled detector output pulses. As flux rates increase, the ability to implement such a detection algorithm becomes difficult within a digital processor that may reside within a field-programmable gate array (FPGA). Systems have been developed and implemented to both characterize gigahertz bandwidth single-photon detectors, as well as process photon count signals at rates into gigaphotons per second in order to implement communications links at SCPPM (serial concatenated pulse position modulation) encoded data rates exceeding 100 megabits per second with efficiencies greater than two bits per detected photon. A hardware edge-detection algorithm and corresponding signal combining and deserialization hardware were developed to meet these requirements at sample rates up to 10 GHz. The photon discriminator deserializer hardware board accepts four inputs, which allows for the ability to take inputs from a quadphoton counting detector, to support requirements for optical tracking with a reduced number of hardware components. The four inputs are hardware leading-edge detected independently. After leading-edge detection, the resultant samples are ORed together prior to deserialization. The deserialization is performed to reduce the rate at which data is passed to a digital signal processor, perhaps residing within an FPGA. The hardware implements four separate analog inputs that are connected through RF connectors. Each analog input is fed to a high-speed 1

  7. AEROSOL NUCLEATION AND GROWTH DURING LAMINAR TUBE FLOW: MAXIMUM SATURATIONS AND NUCLEATION RATES. (R827354C008)

    Science.gov (United States)

    An approximate method of estimating the maximum saturation, the nucleation rate, and the total number nucleated per second during the laminar flow of a hot vapour–gas mixture along a tube with cold walls is described. The basis of the approach is that the temperature an...

  8. Development of amplifier and shaper for high-rate MWPC

    International Nuclear Information System (INIS)

    Kamiji, Ichinori; Nanjo, Hajime; Kawasaki, Naoki; Maeda, Yosuke; Naito, Daichi; Seki, Shigeto; Nakagiri, Kota; Sasao, Noboru; Nomura, Tadashi

    2015-01-01

    A multi-wire proportional chamber (MWPC) will be used as an in-beam charged particle detector for the J-PARC E14 (KOTO) experiment. The maximum counting rate of the MWPC is expected to be up to 1 MHz per channel due to the high neutron and photon flux, expected to be 1 GHz for the 30 x 30 cm"2 area. An amplifier to cope with such high counting-rate is required. We developed a prototype of such amplifier, which has a charge preamplifier with the integration time of 3 ns and a pulse shaping part with three pole-zero cancellation networks. The shaper reduced the characteristic long tail lasting tens of microseconds in the signal of MWPC to 150 ns. Its performance has been tested by using a single-channel MWPC which has almost the same geometrical parameters as the MWPC to be installed in the KOTO experiment. (author)

  9. 22 CFR 201.67 - Maximum freight charges.

    Science.gov (United States)

    2010-04-01

    ..., commodity rate classification, quantity, vessel flag category (U.S.-or foreign-flag), choice of ports, and... the United States. (2) Maximum charter rates. (i) USAID will not finance ocean freight under any... owner(s). (4) Maximum liner rates. USAID will not finance ocean freight for a cargo liner shipment at a...

  10. Material screening with HPGe counting station for PandaX experiment

    International Nuclear Information System (INIS)

    Wang, X.; Chen, X.; Fu, C.; Ji, X.; Liu, X.; Xie, P.; Zhang, T.; Mao, Y.; Wang, S.; Wang, H.

    2016-01-01

    A gamma counting station based on high-purity germanium (HPGe) detector was set up for the material screening of the PandaX dark matter experiments in the China Jinping Underground Laboratory. Low background gamma rate of 2.6 counts/min within the energy range of 20 to 2700 keV is achieved due to the well-designed passive shield. The sentivities of the HPGe detetector reach mBq/kg level for isotopes like K, U, Th, and even better for Co and Cs, resulted from the low-background rate and the high relative detection efficiency of 175%. The structure and performance of the counting station are described in this article. Detailed counting results for the radioactivity in materials used by the PandaX dark-matter experiment are presented. The upgrading plan of the counting station is also discussed.

  11. Preamplifier development for high count-rate, large dynamic range readout of inorganic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Keshelashvili, Irakli; Erni, Werner; Steinacher, Michael; Krusche, Bernd; Collaboration: PANDA-Collaboration

    2013-07-01

    Electromagnetic calorimeter are central component of many experiments in nuclear and particle physics. Modern ''trigger less'' detectors run with very high count-rates, require good time and energy resolution, and large dynamic range. In addition photosensors and preamplifiers must work in hostile environments (magnetic fields). Due to later constraints mainly Avalanche Photo Diodes (APD's), Vacuum Photo Triodes (VPT's), and Vacuum Photo Tetrodes (VPTT's) are used. A disadvantage is their low gain which together with other requirements is a challenge for the preamplifier design. Our group has developed special Low Noise / Low Power (LNP) preamplifier for this purpose. They will be used to equip PANDA EMC forward end-cap (dynamic range 15'000, rate 1MHz), where the PWO II crystals and preamplifier have to run in an environment cooled down to -25{sup o}C. Further application is the upgrade of the Crystal Barrel detector at the Bonn ELSA accelerator with APD readout for which special temperature comparison of the APD gain and good time resolution is necessary. Development and all test procedures after the mass production done by our group during past several years in Basel University will be reported.

  12. Upper limits from counting experiments with multiple pipelines

    International Nuclear Information System (INIS)

    Sutton, Patrick J

    2009-01-01

    In counting experiments, one can set an upper limit on the rate of a Poisson process based on a count of the number of events observed due to the process. In some experiments, one makes several counts of the number of events, using different instruments, different event detection algorithms or observations over multiple time intervals. We demonstrate how to generalize the classical frequentist upper limit calculation to the case where multiple counts of events are made over one or more time intervals using several (not necessarily independent) procedures. We show how different choices of the rank ordering of possible outcomes in the space of counts correspond to applying different levels of significance to the various measurements. We propose an ordering that is matched to the sensitivity of the different measurement procedures and show that in typical cases it gives stronger upper limits than other choices. As an example, we show how this method can be applied to searches for gravitational-wave bursts, where multiple burst-detection algorithms analyse the same data set, and demonstrate how a single combined upper limit can be set on the gravitational-wave burst rate.

  13. The impact of preapheresis white blood cell count on autologous peripheral blood stem cell collection efficiency and HSC infusion side effect rate.

    Science.gov (United States)

    Sakashita, Araci M; Kondo, Andrea T; Yokoyama, Ana Paula H; Lira, Sanny M C; Bub, Carolina B; Souza, Aline M; Cipolletta, Andrea N F; Alvarez, Kelen C; Hamerschlak, Nelson; Kutner, Jose M; Chiattone, Carlos S

    2018-01-19

    Autologous peripheral blood hematopoietic stem cell (PBSC) collection efficiency (CE) is reportedly affected by the patient's blood properties; however, studies to identify factors correlated with CE have shown inconsistent results. Additionally, variables such as stem cell graft granulocyte content and patient age, sex, and underlying disease, may be associated with hematopietic stem cell (HSC) infusion-related adverse reactions. In this study, we evaluated the correlation of preleukapheresis PB granulocyte count and PBSC harvest variables with CD34 + collection yield and efficiency, and thawed HSC infusion side effect occurrence. We evaluated data from 361 patients who had undergone autologous PBSC transplant. Large volume leukapheresis was the method for PBSC collection. Complete Blood Count and CD34 + cell enumeration were performed in the preapheresis PB and the apheresis product sample. The PBSC grafts were submitted to non-controlled rate freezing after addition of 5% DMSO plus 6% hidroxyethylstarch as a cryoprotectant solution. The cryopreserved graft was thawed in a 37°C water bath and then infused without further manipulation. The CD34 + yield was associated with preapheresis PB CD34 + count and immature granulocyte count. The PBSC CE was negatively correlated with preapheresis white blood cell (WBC), immature granulocyte and granulocyte count. The leukapheresis product total nucleated cell (TNC) and granulocyte content was correlated with the thawed graft infusion side effect occurrence. This study has shown that preapheresis PB WBC and granulocyte counts were associated with leukapheresis CE. Additionally, the leukapheresis product TNC and granulocyte content was correlated with thawed graft infusion side effect occurrence. © 2018 Wiley Periodicals, Inc.

  14. Metal ion levels and lymphocyte counts

    DEFF Research Database (Denmark)

    Penny, Jeannette Ø; Varmarken, Jens-Erik; Ovesen, Ole

    2013-01-01

    BACKGROUND AND PURPOSE: Wear particles from metal-on-metal arthroplasties are under suspicion for adverse effects both locally and systemically, and the DePuy ASR Hip Resurfacing System (RHA) has above-average failure rates. We compared lymphocyte counts in RHA and total hip arthroplasty (THA) an....../ppb. INTERPRETATION: Circulating T-lymphocyte levels may decline after surgery, regardless of implant type. Metal ions-particularly cobalt-may have a general depressive effect on T- and B-lymphocyte levels. Registered with ClinicalTrials.gov under # NCT01113762.......BACKGROUND AND PURPOSE: Wear particles from metal-on-metal arthroplasties are under suspicion for adverse effects both locally and systemically, and the DePuy ASR Hip Resurfacing System (RHA) has above-average failure rates. We compared lymphocyte counts in RHA and total hip arthroplasty (THA....... RESULTS: The T-lymphocyte counts for both implant types declined over the 2-year period. This decline was statistically significant for CD3(+)CD8(+) in the THA group, with a regression coefficient of -0.04 × 10(9)cells/year (95% CI: -0.08 to -0.01). Regression analysis indicated a depressive effect...

  15. Antral follicle counts are strongly associated with live-birth rates after assisted reproduction, with superior treatment outcome in women with polycystic ovaries.

    Science.gov (United States)

    Holte, Jan; Brodin, Thomas; Berglund, Lars; Hadziosmanovic, Nermin; Olovsson, Matts; Bergh, Torbjörn

    2011-09-01

    To evaluate the association of antral follicle count (AFC) with in vitro fertilization/intracytoplasmic sperm injection (IVF-ICSI) outcome in a large unselected cohort of patients covering the entire range of AFC. Prospective observational study. University-affiliated private infertility center. 2,092 women undergoing 4,308 IVF-ICSI cycles. AFC analyzed for associations with treatment outcome and statistically adjusted for repeated treatments and age. Pregnancy rate, live-birth rate, and stimulation outcome parameters. The AFC was log-normally distributed. Pregnancy rates and live-birth rates were positively associated with AFC in a log-linear way, leveling out above AFC ∼30. Treatment outcome was superior among women with polycystic ovaries, independent from ovulatory status. The findings were significant also after adjustment for age and number of oocytes retrieved. Pregnancy and live-birth rates are log-linearly related to AFC. Polycystic ovaries, most often excluded from studies on ovarian reserve, fit as one extreme in the spectrum of AFC; a low count constitutes the other extreme, with the lowest ovarian reserve and poor treatment outcome. The findings remained statistically significant also after adjustment for the number of oocytes retrieved, suggesting this measure of ovarian reserve comprises information on oocyte quality and not only quantity. Copyright © 2011 American Society for Reproductive Medicine. Published by Elsevier Inc. All rights reserved.

  16. Accuracy of maximum likelihood estimates of a two-state model in single-molecule FRET

    Energy Technology Data Exchange (ETDEWEB)

    Gopich, Irina V. [Laboratory of Chemical Physics, National Institute of Diabetes and Digestive and Kidney Diseases, National Institutes of Health, Bethesda, Maryland 20892 (United States)

    2015-01-21

    Photon sequences from single-molecule Förster resonance energy transfer (FRET) experiments can be analyzed using a maximum likelihood method. Parameters of the underlying kinetic model (FRET efficiencies of the states and transition rates between conformational states) are obtained by maximizing the appropriate likelihood function. In addition, the errors (uncertainties) of the extracted parameters can be obtained from the curvature of the likelihood function at the maximum. We study the standard deviations of the parameters of a two-state model obtained from photon sequences with recorded colors and arrival times. The standard deviations can be obtained analytically in a special case when the FRET efficiencies of the states are 0 and 1 and in the limiting cases of fast and slow conformational dynamics. These results are compared with the results of numerical simulations. The accuracy and, therefore, the ability to predict model parameters depend on how fast the transition rates are compared to the photon count rate. In the limit of slow transitions, the key parameters that determine the accuracy are the number of transitions between the states and the number of independent photon sequences. In the fast transition limit, the accuracy is determined by the small fraction of photons that are correlated with their neighbors. The relative standard deviation of the relaxation rate has a “chevron” shape as a function of the transition rate in the log-log scale. The location of the minimum of this function dramatically depends on how well the FRET efficiencies of the states are separated.

  17. CERNDxCTA counting mode chip

    International Nuclear Information System (INIS)

    Moraes, D.; Kaplon, J.; Nygard, E.

    2008-01-01

    This ASIC is a counting mode front-end electronic optimized for the readout of CdZnTe/CdTe and silicon sensors, for possible use in applications where the flux of ionizing radiation is high. The chip is implemented in 0.25 μm CMOS technology. The circuit comprises 128 channels equipped with a transimpedance amplifier followed by a gain shaper stage with 21 ns peaking time, two discriminators and two 18-bit counters. The channel architecture is optimized for the detector characteristics in order to achieve the best energy resolution at counting rates of up to 5 M counts/second. The amplifier shows a linear sensitivity of 118 mV/fC and an equivalent noise charge of about 711 e - , for a detector capacitance of 5 pF. Complete evaluation of the circuit is presented using electronic pulses and pixel detectors

  18. Pitch Counts in Youth Baseball and Softball: A Historical Review.

    Science.gov (United States)

    Feeley, Brian T; Schisel, Jessica; Agel, Julie

    2018-07-01

    Pitching injuries are getting increased attention in the mass media. Many references are made to pitch counts and the role they play in injury prevention. The original purpose of regulating the pitch count in youth baseball was to reduce injury and fatigue to pitchers. This article reviews the history and development of the pitch count limit in baseball, the effect it has had on injury, and the evidence regarding injury rates on softball windmill pitching. Literature search through PubMed, mass media, and organizational Web sites through June 2015. Pitch count limits and rest recommendations were introduced in 1996 after a survey of 28 orthopedic surgeons and baseball coaches showed injuries to baseball pitchers' arms were believed to be from the number of pitches thrown. Follow-up research led to revised recommendations with more detailed guidelines in 2006. Since that time, data show a relationship between innings pitched and upper extremity injury, but pitch type has not clearly been shown to affect injury rates. Current surveys of coaches and players show that coaches, parents, and athletes often do not adhere to these guidelines. There are no pitch count guidelines currently available in softball. The increase in participation in youth baseball and softball with an emphasis on early sport specialization in youth sports activities suggests that there will continue to be a rise in injury rates to young throwers. The published pitch counts are likely to positively affect injury rates but must be adhered to by athletes, coaches, and parents.

  19. Standardization of Ga-68 by coincidence measurements, liquid scintillation counting and 4πγ counting

    International Nuclear Information System (INIS)

    Roteta, Miguel; Peyres, Virginia; Rodríguez Barquero, Leonor; García-Toraño, Eduardo; Arenillas, Pablo; Balpardo, Christian; Rodrígues, Darío; Llovera, Roberto

    2012-01-01

    The radionuclide 68 Ga is one of the few positron emitters that can be prepared in-house without the use of a cyclotron. It disintegrates to the ground state of 68 Zn partially by positron emission (89.1%) with a maximum energy of 1899.1 keV, and partially by electron capture (10.9%). This nuclide has been standardized in the frame of a cooperation project between the Radionuclide Metrology laboratories from CIEMAT (Spain) and CNEA (Argentina). Measurements involved several techniques: 4πβ−γ coincidences, integral gamma counting and Liquid Scintillation Counting using the triple to double coincidence ratio and the CIEMAT/NIST methods. Given the short half-life of the radionuclide assayed, a direct comparison between results from both laboratories was excluded and a comparison of experimental efficiencies of similar NaI detectors was used instead. - Highlights: ► We standardized the positron emitter Ga-68 in a bilateral cooperation. ► We used several techniques, as coincidence, integral gamma and liquid scintillation. ► An efficiency comparison replaced a direct comparison of reference materials.

  20. Optimization of statistical methods for HpGe gamma-ray spectrometer used in wide count rate ranges

    Energy Technology Data Exchange (ETDEWEB)

    Gervino, G., E-mail: gervino@to.infn.it [UNITO - Università di Torino, Dipartimento di Fisica, Turin (Italy); INFN - Istituto Nazionale di Fisica Nucleare, Sez. Torino, Turin (Italy); Mana, G. [INRIM - Istituto Nazionale di Ricerca Metrologica, Turin (Italy); Palmisano, C. [UNITO - Università di Torino, Dipartimento di Fisica, Turin (Italy); INRIM - Istituto Nazionale di Ricerca Metrologica, Turin (Italy)

    2016-07-11

    The need to perform γ-ray measurements with HpGe detectors is a common technique in many fields such as nuclear physics, radiochemistry, nuclear medicine and neutron activation analysis. The use of HpGe detectors is chosen in situations where isotope identification is needed because of their excellent resolution. Our challenge is to obtain the “best” spectroscopy data possible in every measurement situation. “Best” is a combination of statistical (number of counts) and spectral quality (peak, width and position) over a wide range of counting rates. In this framework, we applied Bayesian methods and the Ellipsoidal Nested Sampling (a multidimensional integration technique) to study the most likely distribution for the shape of HpGe spectra. In treating these experiments, the prior information suggests to model the likelihood function with a product of Poisson distributions. We present the efforts that have been done in order to optimize the statistical methods to HpGe detector outputs with the aim to evaluate to a better order of precision the detector efficiency, the absolute measured activity and the spectra background. Reaching a more precise knowledge of statistical and systematic uncertainties for the measured physical observables is the final goal of this research project.

  1. Point Count Length and Detection of Forest Neotropical Migrant Birds

    Science.gov (United States)

    Deanna K. Dawson; David R. Smith; Chandler S. Robbins

    1995-01-01

    Comparisons of bird abundances among years or among habitats assume that the rates at which birds are detected and counted are constant within species. We use point count data collected in forests of the Mid-Atlantic states to estimate detection probabilities for Neotropical migrant bird species as a function of count length. For some species, significant differences...

  2. A computer simulation used to investigate optimization in low level counting

    International Nuclear Information System (INIS)

    Brown, R.C.; Kephart, G.S.

    1984-01-01

    The differential form of the interval distribution for randomly spaced events such as radioactive decay is represented as dP/sub t/=ae - /supat/dt, the Poisson distribution. As applied to radioactive decay, this states that the probability (dP/sub t/) of the duration of a particular interval (elapsed time between counts) will be between t and t+dt as a function of the count rate (a). Thus a logarithmic transformation of this probability distribution results in a linear function whose slope and intercept are defined by the count rate. The effort expended in defining the interval distribution of a given radiation measurement equates in the laboratory to measuring and accumulating discrete time intervals between events rather than the usual approach of counting events per unit time. It follows from basic information theory that this greater effort should result in an improved statistical confidence in determinations of the ''true'' count rate (a). Using a random number generator as an analog of the discrete decay event, the authors have devised a Monte Carlo approach to investigate application of the above theory to the low level counting situation. This investigative approach is well suited to sensitivity analyses such that any constraints on proposed optimization techniques can be well defined prior to introducing these methods into the counting requirements in the laboratory

  3. Maximum permissible continuous release rates of phosphorus-32 and sulphur-35 to atmosphere in a milk producing area

    Energy Technology Data Exchange (ETDEWEB)

    Bryant, P M

    1963-01-01

    A method is given for calculating, for design purposes, the maximum permissible continuous release rates of phosphorus-32 and sulphur-35 to atmosphere with respect to milk contamination. In the absence of authoritative advice from the Medical Research Council, provisional working levels for the concentration of phosphorus-32 and sulphur-35 in milk are derived, and details are given of the agricultural assumptions involved in the calculation of the relationship between the amount of the nuclide deposited on grassland and that to be found in milk. The agricultural and meteorological conditions assumed are applicable as an annual average to England and Wales. The results (in mc/day) for phosphorus-32 and sulphur-35 for a number of stack heights and distances are shown graphically; typical values, quoted in a table, include 20 mc/day of phosphorus-32 and 30 mc/day of sulfur-35 as the maximum permissible continuous release rates with respect to ground level releases at a distance of 200 metres from pastureland.

  4. On the stability of performance of NaI(Tl) scintillation spectrometer with FEhU-49 photomultiplier at high counting rates

    International Nuclear Information System (INIS)

    Belousov, A.S.; Vazdik, Ya.A.; Malinovskij, E.N.; Rusakov, S.V.; Solov'ev, Yu.V.; Fomenko, A.M.; Sharejko, P.N.

    1986-01-01

    The dependence of instability in NaI(Tl)-spectrometer characteristics on the instability of photomultiplier (PM), the multiplication factor of which grows with an increase in counting rate, is determined. A simple way to stabilize PM gain factor with an accuracy up to 1.7% is suggested, which consists in stabilization of voltage in two terminal dynodes of PM and photocathode illumination by an auxillary light source

  5. Scintillation counter, maximum gamma aspect

    International Nuclear Information System (INIS)

    Thumim, A.D.

    1975-01-01

    A scintillation counter, particularly for counting gamma ray photons, includes a massive lead radiation shield surrounding a sample-receiving zone. The shield is disassembleable into a plurality of segments to allow facile installation and removal of a photomultiplier tube assembly, the segments being so constructed as to prevent straight-line access of external radiation through the shield into radiation-responsive areas. Provisions are made for accurately aligning the photomultiplier tube with respect to one or more sample-transmitting bores extending through the shield to the sample receiving zone. A sample elevator, used in transporting samples into the zone, is designed to provide a maximum gamma-receiving aspect to maximize the gamma detecting efficiency. (U.S.)

  6. Distribution of ion space charge in the volume of a multiwire proportional chamber at high counting rates

    International Nuclear Information System (INIS)

    Dmitriev, G.D.; Frumkin, I.B.

    1990-01-01

    The accumulation of ion space charge in the drift gap of a MWPC, separated from the amplification gap by a cathode grid, is studied. A significant dependence of the cathode grid transparency to ions on the counting rate is observed. It is shown that the decrease of ion transparency with beam intensity is defined both by the parameters and the mode of operation of the MWPC as well as by the gas mixture chosen. On the basis of suggested explanation of this effect a formula is obtained, that proved to be in good agreement with the experimental results. (orig.)

  7. Relationship between γ detection dead-time and count correction factor

    International Nuclear Information System (INIS)

    Wu Huailong; Zhang Jianhua; Chu Chengsheng; Hu Guangchun; Zhang Changfan; Hu Gen; Gong Jian; Tian Dongfeng

    2015-01-01

    The relationship between dead-time and count correction factor was investigated by using interference source for purpose of high γ activity measurement. The count rates maintain several 10 s"-"l with γ energy of 0.3-1.3 MeV for 10"4-10"5 Bq radioactive source. It is proved that the relationship between count loss and dead-time is unconcerned at various energy and various count intensities. The same correction formula can be used for any nuclide measurement. (authors)

  8. Correlation between discharged worms and fecal egg counts in human clonorchiasis.

    Directory of Open Access Journals (Sweden)

    Jae-Hwan Kim

    2011-10-01

    Full Text Available BACKGROUND: Stool examination by counting eggs per gram of feces (EPGs is the best method to estimate worm burden of Clonorchis sinensis in infected humans. The present study investigated a correlation between EPGs and worm burden in human clonorchiasis. METHODS AND FINDINGS: A total of 60 residents, 50 egg-positive and 10 egg-negative, in Sancheong-gun, Korea, participated in this worm collection trial in 2006-2009. They were diagnosed by egg positivity in feces using the Kato-Katz method. After administration of praziquantel, they were purged with cathartics on the next day, and then discharged adult worms were collected from their feces. Their EPGs ranged from 0 to 65,544. Adult worms of C. sinensis were collected from 17 egg-positive cases, and the number of worms ranged from 1 to 114 in each individual. A positive correlation between EPGs and numbers of worms was demonstrated (r = 0.681, P<0.001. Worm recovery rates were 9.7% in cases of EPGs 1-1,000 and 73.7% in those of EPGs over 1,000. No worms were detected from egg-negative subjects. Maximum egg count per worm per day was roughly estimated 3,770 in a subject with EPGs 2,664 and 106 collected worms. CONCLUSIONS: The numbers of the worms are significantly correlated with the egg counts in human clonorchiasis. It is estimated that at least 110 worms are infected in a human body with EPGs around 3,000, and egg productivity of a worm per day is around 4,000.

  9. Maximum type I error rate inflation from sample size reassessment when investigators are blind to treatment labels.

    Science.gov (United States)

    Żebrowska, Magdalena; Posch, Martin; Magirr, Dominic

    2016-05-30

    Consider a parallel group trial for the comparison of an experimental treatment to a control, where the second-stage sample size may depend on the blinded primary endpoint data as well as on additional blinded data from a secondary endpoint. For the setting of normally distributed endpoints, we demonstrate that this may lead to an inflation of the type I error rate if the null hypothesis holds for the primary but not the secondary endpoint. We derive upper bounds for the inflation of the type I error rate, both for trials that employ random allocation and for those that use block randomization. We illustrate the worst-case sample size reassessment rule in a case study. For both randomization strategies, the maximum type I error rate increases with the effect size in the secondary endpoint and the correlation between endpoints. The maximum inflation increases with smaller block sizes if information on the block size is used in the reassessment rule. Based on our findings, we do not question the well-established use of blinded sample size reassessment methods with nuisance parameter estimates computed from the blinded interim data of the primary endpoint. However, we demonstrate that the type I error rate control of these methods relies on the application of specific, binding, pre-planned and fully algorithmic sample size reassessment rules and does not extend to general or unplanned sample size adjustments based on blinded data. © 2015 The Authors. Statistics in Medicine Published by John Wiley & Sons Ltd. © 2015 The Authors. Statistics in Medicine Published by John Wiley & Sons Ltd.

  10. Experimental procedure for the determination of counting efficiency and sampling flow rate of a grab-sampling working level meter

    International Nuclear Information System (INIS)

    Grenier, M.; Bigu, J.

    1982-07-01

    The calibration procedures used for a working level meter (WLM) of the grab-sampling type are presented in detail. The WLM tested is a Pylon WL-1000C working level meter and it was calibrated for radon/thoron daughter counting efficiency (E), for sampling pump flow rate (Q) and other variables of interest. For the instrument calibrated at the Elliot Lake Laboratory, E was 0.22 +- 0.01 while Q was 4.50 +- 0.01 L/min

  11. Performance of n-γ pulse-shape discrimination with simple pile-up rejection at high γ-ray count rates

    International Nuclear Information System (INIS)

    Okuda, T.; Yamazaki, H.; Kawabata, M.; Kasagi, J.; Harada, H.

    1999-01-01

    The performance of n-γ pulse-shape discrimination for a liquid scintillation detector has been investigated for γ-ray count rates up to 50 kcps. A method in which the ratio of the total to partial charge in the anode pulse is directly measured has shown much improved quality of the pulse-shape discrimination when pile-up events are rejected; it can discriminate neutron events of 50 cps from γ-ray events of 29 kcps. The method with simple pile-up rejection has the advantage that only general purpose electronics are required

  12. Non-Poisson counting statistics of a hybrid G-M counter dead time model

    International Nuclear Information System (INIS)

    Lee, Sang Hoon; Jae, Moosung; Gardner, Robin P.

    2007-01-01

    The counting statistics of a G-M counter with a considerable dead time event rate deviates from Poisson statistics. Important characteristics such as observed counting rates as a function true counting rates, variances and interval distributions were analyzed for three dead time models, non-paralyzable, paralyzable and hybrid, with the help of GMSIM, a Monte Carlo dead time effect simulator. The simulation results showed good agreements with the models in observed counting rates and variances. It was found through GMSIM simulations that the interval distribution for the hybrid model showed three distinctive regions, a complete cutoff region for the duration of the total dead time, a degraded exponential and an enhanced exponential regions. By measuring the cutoff and the duration of degraded exponential from the pulse interval distribution, it is possible to evaluate the two dead times in the hybrid model

  13. EcoCount

    Directory of Open Access Journals (Sweden)

    Phillip P. Allen

    2014-05-01

    Full Text Available Techniques that analyze biological remains from sediment sequences for environmental reconstructions are well established and widely used. Yet, identifying, counting, and recording biological evidence such as pollen grains remain a highly skilled, demanding, and time-consuming task. Standard procedure requires the classification and recording of between 300 and 500 pollen grains from each representative sample. Recording the data from a pollen count requires significant effort and focused resources from the palynologist. However, when an adaptation to the recording procedure is utilized, efficiency and time economy improve. We describe EcoCount, which represents a development in environmental data recording procedure. EcoCount is a voice activated fully customizable digital count sheet that allows the investigator to continuously interact with a field of view during the data recording. Continuous viewing allows the palynologist the opportunity to remain engaged with the essential task, identification, for longer, making pollen counting more efficient and economical. EcoCount is a versatile software package that can be used to record a variety of environmental evidence and can be installed onto different computer platforms, making the adoption by users and laboratories simple and inexpensive. The user-friendly format of EcoCount allows any novice to be competent and functional in a very short time.

  14. 15 Mcps photon-counting X-ray computed tomography system using a ZnO-MPPC detector and its application to gadolinium imaging

    Energy Technology Data Exchange (ETDEWEB)

    Sato, Eiichi, E-mail: dresato@iwate-med.ac.jp [Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694 (Japan); Sugimura, Shigeaki [Tokyo Denpa Co. Ltd., 82-5 Ueno, Ichinohe, Iwate 028-5321 (Japan); Endo, Haruyuki [Iwate Industrial Research Insutitute 3, 3-35-2 Shinden, Iioka, Morioka, Iwate 020-0852 (Japan); Oda, Yasuyuki [Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694 (Japan); Abudurexiti, Abulajiang [Faculty of Software and Information Science, Iwate Prefectural University, 152-52 Sugo, Takizawa, Iwate 020-0193 (Japan); Hagiwara, Osahiko; Osawa, Akihiro; Matsukiyo, Hiroshi; Enomoto, Toshiyuki; Watanabe, Manabu; Kusachi, Shinya [3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro-ku, Tokyo 153-8515 (Japan); Sato, Shigehiro [Department of Microbiology, School of Medicine, Iwate Medical University, 19-1 Uchimaru, Morioka, Iwate 020-0023 (Japan); Ogawa, Akira [Department of Neurosurgery, School of Medicine, Iwate Medical University, 19-1 Uchimaru, Morioka, Iwate 020-0023 (Japan); Onagawa, Jun [Department of Electronics, Faculty of Engineering, Tohoku Gakuin University, 1-13-1 Chuo, Tagajo, Miyagi 985-8537 (Japan)

    2012-01-15

    15 Mcps photon-counting X-ray computed tomography (CT) system is a first-generation type and consists of an X-ray generator, a turntable, a translation stage, a two-stage controller, a detector consisting of a 2 mm-thick zinc-oxide (ZnO) single-crystal scintillator and an MPPC (multipixel photon counter) module, a counter card (CC), and a personal computer (PC). High-speed photon counting was carried out using the detector in the X-ray CT system. The maximum count rate was 15 Mcps (mega counts per second) at a tube voltage of 100 kV and a tube current of 1.95 mA. Tomography is accomplished by repeated translations and rotations of an object, and projection curves of the object are obtained by the translation. The pulses of the event signal from the module are counted by the CC in conjunction with the PC. The minimum exposure time for obtaining a tomogram was 15 min, and photon-counting CT was accomplished using gadolinium-based contrast media. - Highlights: Black-Right-Pointing-Pointer We developed a first-generation 15 Mcps photon-counting X-ray computed tomography (CT) system. Black-Right-Pointing-Pointer High-speed photon counting was carried out using a zinc-oxide (ZnO) single-crystal scintillator and an MPPC (multipixel photon counter) module in the X-ray CT system. Black-Right-Pointing-Pointer Tomography is accomplished by repeated translations and rotations of an object. Black-Right-Pointing-Pointer The minimum exposure time for obtaining a tomogram was 15 min. Black-Right-Pointing-Pointer The photon-counting CT was accomplished using gadolinium-based contrast media.

  15. RBC count

    Science.gov (United States)

    ... by kidney disease) RBC destruction ( hemolysis ) due to transfusion, blood vessel injury, or other cause Leukemia Malnutrition Bone ... slight risk any time the skin is broken) Alternative Names Erythrocyte count; Red blood cell count; Anemia - RBC count Images Blood test ...

  16. Tutorial on X-ray photon counting detector characterization.

    Science.gov (United States)

    Ren, Liqiang; Zheng, Bin; Liu, Hong

    2018-01-01

    Recent advances in photon counting detection technology have led to significant research interest in X-ray imaging. As a tutorial level review, this paper covers a wide range of aspects related to X-ray photon counting detector characterization. The tutorial begins with a detailed description of the working principle and operating modes of a pixelated X-ray photon counting detector with basic architecture and detection mechanism. Currently available methods and techniques for charactering major aspects including energy response, noise floor, energy resolution, count rate performance (detector efficiency), and charge sharing effect of photon counting detectors are comprehensively reviewed. Other characterization aspects such as point spread function (PSF), line spread function (LSF), contrast transfer function (CTF), modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), bias voltage, radiation damage, and polarization effect are also remarked. A cadmium telluride (CdTe) pixelated photon counting detector is employed for part of the characterization demonstration and the results are presented. This review can serve as a tutorial for X-ray imaging researchers and investigators to understand, operate, characterize, and optimize photon counting detectors for a variety of applications.

  17. Fast counting electronics for neutron coincidence counting

    International Nuclear Information System (INIS)

    Swansen, J.E.

    1987-01-01

    This patent describes a high speed circuit for accurate neutron coincidence counting comprising: neutron detecting means for providing an above-threshold signal upon neutron detection; amplifying means inputted by the neutron detecting means for providing a pulse output having a pulse width of about 0.5 microseconds upon the input of each above threshold signal; digital processing means inputted by the pulse output of the amplifying means for generating a pulse responsive to each input pulse from the amplifying means and having a pulse width of about 50 nanoseconds effective for processing an expected neutron event rate of about 1 Mpps: pulse stretching means inputted by the digital processing means for producing a pulse having a pulse width of several milliseconds for each pulse received form the digital processing means; visual indicating means inputted by the pulse stretching means for producing a visual output for each pulse received from the digital processing means; and derandomizing means effective to receive the 50 ns neutron event pulses from the digital processing means for storage at a rate up to the neutron event rate of 1 Mpps and having first counter means for storing the input neutron event pulses

  18. Outlier identification procedures for contingency tables using maximum likelihood and $L_1$ estimates

    NARCIS (Netherlands)

    Kuhnt, S.

    2004-01-01

    Observed cell counts in contingency tables are perceived as outliers if they have low probability under an anticipated loglinear Poisson model. New procedures for the identification of such outliers are derived using the classical maximum likelihood estimator and an estimator based on the L1 norm.

  19. CERN_DxCTA counting mode chip

    CERN Document Server

    Moraes, D; Nygård, E

    2008-01-01

    This ASIC is a counting mode front-end electronic optimized for the readout of CdZnTe/CdTe and silicon sensors, for possible use in applications where the flux of ionizing radiation is high. The chip is implemented in 0.25 μm CMOS technology. The circuit comprises 128 channels equipped with a transimpedance amplifier followed by a gain shaper stage with 21 ns peaking time, two discriminators and two 18-bit counters. The channel architecture is optimized for the detector characteristics in order to achieve the best energy resolution at counting rates of up to 5 M counts/second. The amplifier shows a linear sensitivity of 118 mV/fC and an equivalent noise charge of about 711 e−, for a detector capacitance of 5 pF. Complete evaluation of the circuit is presented using electronic pulses and pixel detectors.

  20. Monolithic array of 32 SPAD pixels for single-photon imaging at high frame rates

    International Nuclear Information System (INIS)

    Tisa, Simone; Guerrieri, Fabrizio; Zappa, Franco

    2009-01-01

    We present a single-chip monolithic array of 32 Single-Photon Avalanche Diodes (SPAD) and associated electronics for imaging at high frame rates and high sensitivity. Photodetectors, front-end circuitry and control electronics used to manage the array are monolithically integrated on the same chip in a standard 0.35 μm CMOS high-voltage technology. The array is composed of 32 'smart' pixels working in photon counting mode and functioning in a parallel fashion. Every cell comprises of an integrated SPAD photodetector, a novel quenching circuit named as Variable Load Quenching Circuit (VLQC), counting electronics and a buffer memory. Proper ancillary electronics that perform the arbitration of photon counts between two consecutive frames is integrated as well. Thanks to the presence of in-pixel memory registers, the inter-frame dead time between subsequent frames is limited to few nanoseconds. Since integration and download are performed simultaneously and the array can be addressed like a standard digital memory, the achievable maximum frame rate is very high in the order of hundreds of thousands of frame/s.

  1. Analysis of dental caries using generalized linear and count regression models

    Directory of Open Access Journals (Sweden)

    Javali M. Phil

    2013-11-01

    Full Text Available Generalized linear models (GLM are generalization of linear regression models, which allow fitting regression models to response data in all the sciences especially medical and dental sciences that follow a general exponential family. These are flexible and widely used class of such models that can accommodate response variables. Count data are frequently characterized by overdispersion and excess zeros. Zero-inflated count models provide a parsimonious yet powerful way to model this type of situation. Such models assume that the data are a mixture of two separate data generation processes: one generates only zeros, and the other is either a Poisson or a negative binomial data-generating process. Zero inflated count regression models such as the zero-inflated Poisson (ZIP, zero-inflated negative binomial (ZINB regression models have been used to handle dental caries count data with many zeros. We present an evaluation framework to the suitability of applying the GLM, Poisson, NB, ZIP and ZINB to dental caries data set where the count data may exhibit evidence of many zeros and over-dispersion. Estimation of the model parameters using the method of maximum likelihood is provided. Based on the Vuong test statistic and the goodness of fit measure for dental caries data, the NB and ZINB regression models perform better than other count regression models.

  2. Microbiological assessment of house and imported bottled water by comparison of bacterial endotoxin concentration, heterotrophic plate count, and fecal coliform count.

    Science.gov (United States)

    Reyes, Mayra I; Pérez, Cynthia M; Negrón, Edna L

    2008-03-01

    Consumers increasingly use bottled water and home water treatment systems to avoid direct tap water. According to the International Bottled Water Association (IBWA), an industry trade group, 5 billion gallons of bottled water were consumed by North Americans in 2001. The principal aim of this study was to assess the microbial quality of in-house and imported bottled water for human consumption, by measurement and comparison of the concentration of bacterial endotoxin and standard cultivable methods of indicator microorganisms, specifically, heterotrophic and fecal coliform plate counts. A total of 21 brands of commercial bottled water, consisting of 10 imported and 11 in-house brands, selected at random from 96 brands that are consumed in Puerto Rico, were tested at three different time intervals. The Standard Limulus Amebocyte Lysate test, gel clot method, was used to measure the endotoxin concentrations. The minimum endotoxin concentration in 63 water samples was less than 0.0625 EU/mL, while the maximum was 32 EU/mL. The minimum bacterial count showed no growth, while the maximum was 7,500 CFU/mL. Bacterial isolates like P. fluorescens, Corynebacterium sp. J-K, S. paucimobilis, P. versicularis, A. baumannii, P. chlororaphis, F. indologenes, A. faecalis and P. cepacia were identified. Repeated measures analysis of variance demonstrated that endotoxin concentration did not change over time, while there was a statistically significant (p bacterial count over time. In addition, multiple linear regression analysis demonstrated that a unit change in the concentration of endotoxin across time was associated with a significant (p bacterial growth was not detected in some water samples, endotoxin was present. Measurement of Gram-negative bacterial endotoxins is one of the methods that have been suggested as a rapid way of determining bacteriological water quality.

  3. 2Kx2K resolution element photon counting MCP sensor with >200 kHz event rate capability

    CERN Document Server

    Vallerga, J V

    2000-01-01

    Siegmund Scientific undertook a NASA Small Business Innovative Research (SBIR) contract to develop a versatile, high-performance photon (or particle) counting detector combining recent technical advances in all aspects of Microchannel Plate (MCP) detector development in a low cost, commercially viable package that can support a variety of applications. The detector concept consists of a set of MCPs whose output electron pulses are read out with a crossed delay line (XDL) anode and associated high-speed event encoding electronics. The delay line anode allows high-resolution photon event centroiding at very high event rates and can be scaled to large formats (>40 mm) while maintaining good linearity and high temporal stability. The optimal sensitivity wavelength range is determined by the choice of opaque photocathodes. Specific achievements included: spatial resolution of 200 000 events s sup - sup 1; local rates of >100 events s sup - sup 1 per resolution element; event timing of <1 ns; and low background ...

  4. Categorical counting.

    Science.gov (United States)

    Fetterman, J Gregor; Killeen, P Richard

    2010-09-01

    Pigeons pecked on three keys, responses to one of which could be reinforced after a few pecks, to a second key after a somewhat larger number of pecks, and to a third key after the maximum pecking requirement. The values of the pecking requirements and the proportion of trials ending with reinforcement were varied. Transits among the keys were an orderly function of peck number, and showed approximately proportional changes with changes in the pecking requirements, consistent with Weber's law. Standard deviations of the switch points between successive keys increased more slowly within a condition than across conditions. Changes in reinforcement probability produced changes in the location of the psychometric functions that were consistent with models of timing. Analyses of the number of pecks emitted and the duration of the pecking sequences demonstrated that peck number was the primary determinant of choice, but that passage of time also played some role. We capture the basic results with a standard model of counting, which we qualify to account for the secondary experiments. Copyright 2010 Elsevier B.V. All rights reserved.

  5. Counting statistics in low level radioactivity measurements fluctuating counting efficiency

    International Nuclear Information System (INIS)

    Pazdur, M.F.

    1976-01-01

    A divergence between the probability distribution of the number of nuclear disintegrations and the number of observed counts, caused by counting efficiency fluctuation, is discussed. The negative binominal distribution is proposed to describe the probability distribution of the number of counts, instead of Poisson distribution, which is assumed to hold for the number of nuclear disintegrations only. From actual measurements the r.m.s. amplitude of counting efficiency fluctuation is estimated. Some consequences of counting efficiency fluctuation are investigated and the corresponding formulae are derived: (1) for detection limit as a function of the number of partial measurements and the relative amplitude of counting efficiency fluctuation, and (2) for optimum allocation of the number of partial measurements between sample and background. (author)

  6. Efficient Levenberg-Marquardt minimization of the maximum likelihood estimator for Poisson deviates

    International Nuclear Information System (INIS)

    Laurence, T.; Chromy, B.

    2010-01-01

    Histograms of counted events are Poisson distributed, but are typically fitted without justification using nonlinear least squares fitting. The more appropriate maximum likelihood estimator (MLE) for Poisson distributed data is seldom used. We extend the use of the Levenberg-Marquardt algorithm commonly used for nonlinear least squares minimization for use with the MLE for Poisson distributed data. In so doing, we remove any excuse for not using this more appropriate MLE. We demonstrate the use of the algorithm and the superior performance of the MLE using simulations and experiments in the context of fluorescence lifetime imaging. Scientists commonly form histograms of counted events from their data, and extract parameters by fitting to a specified model. Assuming that the probability of occurrence for each bin is small, event counts in the histogram bins will be distributed according to the Poisson distribution. We develop here an efficient algorithm for fitting event counting histograms using the maximum likelihood estimator (MLE) for Poisson distributed data, rather than the non-linear least squares measure. This algorithm is a simple extension of the common Levenberg-Marquardt (L-M) algorithm, is simple to implement, quick and robust. Fitting using a least squares measure is most common, but it is the maximum likelihood estimator only for Gaussian-distributed data. Non-linear least squares methods may be applied to event counting histograms in cases where the number of events is very large, so that the Poisson distribution is well approximated by a Gaussian. However, it is not easy to satisfy this criterion in practice - which requires a large number of events. It has been well-known for years that least squares procedures lead to biased results when applied to Poisson-distributed data; a recent paper providing extensive characterization of these biases in exponential fitting is given. The more appropriate measure based on the maximum likelihood estimator (MLE

  7. Information loss for 2 × 2 tables with missing cell counts: binomial case

    NARCIS (Netherlands)

    Eisinga, R.N.

    2008-01-01

    We formulate likelihood-based ecological inference for 2 × 2 tables with missing cell counts as an incomplete data problem and study Fisher information loss by comparing estimation from complete and incomplete data. In so doing, we consider maximum-likelihood (ML) estimators of probabilities

  8. Information loss for 2×2 tables with missing cell counts : binomial case

    NARCIS (Netherlands)

    Eisinga, Rob

    2008-01-01

    We formulate likelihood-based ecological inference for 2×2 tables with missing cell counts as an incomplete data problem and study Fisher information loss by comparing estimation from complete and incomplete data. In so doing, we consider maximum-likelihood (ML) estimators of probabilities governed

  9. Complementary frame reconstruction: a low-biased dynamic PET technique for low count density data in projection space

    International Nuclear Information System (INIS)

    Hong, Inki; Cho, Sanghee; Michel, Christian J; Casey, Michael E; Schaefferkoetter, Joshua D

    2014-01-01

    A new data handling method is presented for improving the image noise distribution and reducing bias when reconstructing very short frames from low count dynamic PET acquisition. The new method termed ‘Complementary Frame Reconstruction’ (CFR) involves the indirect formation of a count-limited emission image in a short frame through subtraction of two frames with longer acquisition time, where the short time frame data is excluded from the second long frame data before the reconstruction. This approach can be regarded as an alternative to the AML algorithm recently proposed by Nuyts et al, as a method to reduce the bias for the maximum likelihood expectation maximization (MLEM) reconstruction of count limited data. CFR uses long scan emission data to stabilize the reconstruction and avoids modification of algorithms such as MLEM. The subtraction between two long frame images, naturally allows negative voxel values and significantly reduces bias introduced in the final image. Simulations based on phantom and clinical data were used to evaluate the accuracy of the reconstructed images to represent the true activity distribution. Applicability to determine the arterial input function in human and small animal studies is also explored. In situations with limited count rate, e.g. pediatric applications, gated abdominal, cardiac studies, etc., or when using limited doses of short-lived isotopes such as 15 O-water, the proposed method will likely be preferred over independent frame reconstruction to address bias and noise issues. (paper)

  10. Sperm count as a surrogate endpoint for male fertility control.

    Science.gov (United States)

    Benda, Norbert; Gerlinger, Christoph

    2007-11-30

    When assessing the effectiveness of a hormonal method of fertility control in men, the classical approach used for the assessment of hormonal contraceptives in women, by estimating the pregnancy rate or using a life-table analysis for the time to pregnancy, is difficult to apply in a clinical development program. The main reasons are the dissociation of the treated unit, i.e. the man, and the observed unit, i.e. his female partner, the high variability in the frequency of male intercourse, the logistical cost and ethical concerns related to the monitoring of the trial. A reasonable surrogate endpoint of the definite endpoint time to pregnancy is sperm count. In addition to the avoidance of the mentioned problems, trials that compare different treatments are possible with reasonable sample sizes, and study duration can be shorter. However, current products do not suppress sperm production to 100 per cent in all men and the sperm count is only observed with measurement error. Complete azoospermia might not be necessary in order to achieve an acceptable failure rate compared with other forms of male fertility control. Therefore, the use of sperm count as a surrogate endpoint must rely on the results of a previous trial in which both the definitive- and surrogate-endpoint results were assessed. The paper discusses different estimation functions of the mean pregnancy rate (corresponding to the cumulative hazard) that are based on the results of sperm count trial and a previous trial in which both sperm count and time to pregnancy were assessed, as well as the underlying assumptions. Sample size estimations are given for pregnancy rate estimation with a given precision.

  11. HgCdTe APD-based linear-mode photon counting components and ladar receivers

    Science.gov (United States)

    Jack, Michael; Wehner, Justin; Edwards, John; Chapman, George; Hall, Donald N. B.; Jacobson, Shane M.

    2011-05-01

    Linear mode photon counting (LMPC) provides significant advantages in comparison with Geiger Mode (GM) Photon Counting including absence of after-pulsing, nanosecond pulse to pulse temporal resolution and robust operation in the present of high density obscurants or variable reflectivity objects. For this reason Raytheon has developed and previously reported on unique linear mode photon counting components and modules based on combining advanced APDs and advanced high gain circuits. By using HgCdTe APDs we enable Poisson number preserving photon counting. A metric of photon counting technology is dark count rate and detection probability. In this paper we report on a performance breakthrough resulting from improvement in design, process and readout operation enabling >10x reduction in dark counts rate to ~10,000 cps and >104x reduction in surface dark current enabling long 10 ms integration times. Our analysis of key dark current contributors suggest that substantial further reduction in DCR to ~ 1/sec or less can be achieved by optimizing wavelength, operating voltage and temperature.

  12. Maximum power point tracking

    International Nuclear Information System (INIS)

    Enslin, J.H.R.

    1990-01-01

    A well engineered renewable remote energy system, utilizing the principal of Maximum Power Point Tracking can be m ore cost effective, has a higher reliability and can improve the quality of life in remote areas. This paper reports that a high-efficient power electronic converter, for converting the output voltage of a solar panel, or wind generator, to the required DC battery bus voltage has been realized. The converter is controlled to track the maximum power point of the input source under varying input and output parameters. Maximum power point tracking for relative small systems is achieved by maximization of the output current in a battery charging regulator, using an optimized hill-climbing, inexpensive microprocessor based algorithm. Through practical field measurements it is shown that a minimum input source saving of 15% on 3-5 kWh/day systems can easily be achieved. A total cost saving of at least 10-15% on the capital cost of these systems are achievable for relative small rating Remote Area Power Supply systems. The advantages at larger temperature variations and larger power rated systems are much higher. Other advantages include optimal sizing and system monitor and control

  13. Comparison of least-squares vs. maximum likelihood estimation for standard spectrum technique of β−γ coincidence spectrum analysis

    International Nuclear Information System (INIS)

    Lowrey, Justin D.; Biegalski, Steven R.F.

    2012-01-01

    The spectrum deconvolution analysis tool (SDAT) software code was written and tested at The University of Texas at Austin utilizing the standard spectrum technique to determine activity levels of Xe-131m, Xe-133m, Xe-133, and Xe-135 in β–γ coincidence spectra. SDAT was originally written to utilize the method of least-squares to calculate the activity of each radionuclide component in the spectrum. Recently, maximum likelihood estimation was also incorporated into the SDAT tool. This is a robust statistical technique to determine the parameters that maximize the Poisson distribution likelihood function of the sample data. In this case it is used to parameterize the activity level of each of the radioxenon components in the spectra. A new test dataset was constructed utilizing Xe-131m placed on a Xe-133 background to compare the robustness of the least-squares and maximum likelihood estimation methods for low counting statistics data. The Xe-131m spectra were collected independently from the Xe-133 spectra and added to generate the spectra in the test dataset. The true independent counts of Xe-131m and Xe-133 are known, as they were calculated before the spectra were added together. Spectra with both high and low counting statistics are analyzed. Studies are also performed by analyzing only the 30 keV X-ray region of the β–γ coincidence spectra. Results show that maximum likelihood estimation slightly outperforms least-squares for low counting statistics data.

  14. Improving Bayesian credibility intervals for classifier error rates using maximum entropy empirical priors.

    Science.gov (United States)

    Gustafsson, Mats G; Wallman, Mikael; Wickenberg Bolin, Ulrika; Göransson, Hanna; Fryknäs, M; Andersson, Claes R; Isaksson, Anders

    2010-06-01

    Successful use of classifiers that learn to make decisions from a set of patient examples require robust methods for performance estimation. Recently many promising approaches for determination of an upper bound for the error rate of a single classifier have been reported but the Bayesian credibility interval (CI) obtained from a conventional holdout test still delivers one of the tightest bounds. The conventional Bayesian CI becomes unacceptably large in real world applications where the test set sizes are less than a few hundred. The source of this problem is that fact that the CI is determined exclusively by the result on the test examples. In other words, there is no information at all provided by the uniform prior density distribution employed which reflects complete lack of prior knowledge about the unknown error rate. Therefore, the aim of the study reported here was to study a maximum entropy (ME) based approach to improved prior knowledge and Bayesian CIs, demonstrating its relevance for biomedical research and clinical practice. It is demonstrated how a refined non-uniform prior density distribution can be obtained by means of the ME principle using empirical results from a few designs and tests using non-overlapping sets of examples. Experimental results show that ME based priors improve the CIs when employed to four quite different simulated and two real world data sets. An empirically derived ME prior seems promising for improving the Bayesian CI for the unknown error rate of a designed classifier. Copyright 2010 Elsevier B.V. All rights reserved.

  15. Determining random counts in liquid scintillation counting

    International Nuclear Information System (INIS)

    Horrocks, D.L.

    1979-01-01

    During measurements involving coincidence counting techniques, errors can arise due to the detection of chance or random coincidences in the multiple detectors used. A method and the electronic circuits necessary are here described for eliminating this source of error in liquid scintillation detectors used in coincidence counting. (UK)

  16. Cerenkov counting and Cerenkov-scintillation counting with high refractive index organic liquids using a liquid scintillation counter

    International Nuclear Information System (INIS)

    Wiebe, L.I.; Helus, F.; Maier-Borst, W.

    1978-01-01

    18 F and 14 C radioactivity was measured in methyl salicylate (MS), a high refractive index hybrid Cherenkov-scintillation generating medium, using a liquid scintillation counter. At concentrations of up to 21.4%, in MS, dimethyl sulfoxide (DMSO) quenched 14 C fluorescence, and with a 10-fold excess of DMSO over MS, 18 F count rates were reduced below that for DMSO alone, probably as a result of concentration-independent self-quenching due to 'dark-complex' formation. DMSO in lower concentrations did not reduce the counting efficiency of 18 F in MS. Nitrobenzene was a concentration-dependent quencher for both 14 C and 18 F in MS. Chlorobenzene (CB) and DMSO were both found to be weak Cherenkov generators with 18 F. Counting efficiencies for 18 F in MS, CB, and DMSO were 50.3, 7.8 and 4.3% respectively in the coincidence counting mode, and 58.1, 13.0 and 6.8% in the singles mode. 14 C efficiencies were 14.4 and 22.3% for coincidence and singles respectively, and 15.3 and 42.0% using a modern counter designed for coincidence and single photon counting. The high 14 C and 18 F counting efficiency in MS are discussed with respect to excitation mechanism, on the basis of quench and channels ratios changes observed. It is proposed that MS functions as an efficient Cherenkov-scintillation generator for high-energy beta emitters such as 18 F, and as a low-efficiency scintillator for weak beta emitting radionuclides such as 14 C. (author)

  17. Formal comment on: Myhrvold (2016) Dinosaur metabolism and the allometry of maximum growth rate. PLoS ONE; 11(11): e0163205.

    Science.gov (United States)

    Griebeler, Eva Maria; Werner, Jan

    2018-01-01

    In his 2016 paper, Myhrvold criticized ours from 2014 on maximum growth rates (Gmax, maximum gain in body mass observed within a time unit throughout an individual's ontogeny) and thermoregulation strategies (ectothermy, endothermy) of 17 dinosaurs. In our paper, we showed that Gmax values of similar-sized extant ectothermic and endothermic vertebrates overlap. This strongly questions a correct assignment of a thermoregulation strategy to a dinosaur only based on its Gmax and (adult) body mass (M). Contrary, Gmax separated similar-sized extant reptiles and birds (Sauropsida) and Gmax values of our studied dinosaurs were similar to those seen in extant similar-sized (if necessary scaled-up) fast growing ectothermic reptiles. Myhrvold examined two hypotheses (H1 and H2) regarding our study. However, we did neither infer dinosaurian thermoregulation strategies from group-wide averages (H1) nor were our results based on that Gmax and metabolic rate (MR) are related (H2). In order to assess whether single dinosaurian Gmax values fit to those of extant endotherms (birds) or of ectotherms (reptiles), we already used a method suggested by Myhrvold to avoid H1, and we only discussed pros and cons of a relation between Gmax and MR and did not apply it (H2). We appreciate Myhrvold's efforts in eliminating the correlation between Gmax and M in order to statistically improve vertebrate scaling regressions on maximum gain in body mass. However, we show here that his mass-specific maximum growth rate (kC) replacing Gmax (= MkC) does not model the expected higher mass gain in larger than in smaller species for any set of species. We also comment on, why we considered extant reptiles and birds as reference models for extinct dinosaurs and why we used phylogenetically-informed regression analysis throughout our study. Finally, we question several arguments given in Myhrvold in order to support his results.

  18. Direct photon-counting scintillation detector readout using an SSPM

    International Nuclear Information System (INIS)

    Stapels, Christopher J.; Squillante, Michael R.; Lawrence, William G.; Augustine, Frank L.; Christian, James F.

    2007-01-01

    Gamma-ray detector technologies, capable of providing adequate energy information, use photomultiplier tubes (PMTs) or silicon avalanche photodiodes to detect the light pulse from a scintillation crystal. A new approach to detect the light from scintillation materials is to use an array of small photon counting detectors, or a 'detector-on-a-chip' based on a novel 'Solid-state Photomultiplier' (SSPM) concept. A CMOS SSPM coupled to a scintillation crystal uses an array of CMOS Geiger photodiode (GPD) pixels to collect light and produce a signal proportional to the energy of the radiation. Each pixel acts as a binary photon detector, but the summed output is an analog representation of the total photon intensity. We have successfully fabricated arrays of GPD pixels in a CMOS environment, which makes possible the production of miniaturized arrays integrated with the detector electronics in a small silicon chip. This detector technology allows for a substantial cost reduction while preserving the energy resolution needed for radiological measurements. In this work, we compare designs for the SSPM detector. One pixel design achieves maximum detection efficiency (DE) for 632-nm photons approaching 30% with a room temperature dark count rate (DCR) of less than 1 kHz for a 30-μm-diameter pixel. We characterize after pulsing and optical cross talk and discuss their effects on the performance of the SSPM. For 30-μm diameter, passively quenched CMOS GPD pixels, modeling suggests that a pixel spacing of approximately 90 μm optimizes the SSPM performance with respect to DE and cross talk

  19. Radiation counting statistics

    Energy Technology Data Exchange (ETDEWEB)

    Suh, M. Y.; Jee, K. Y.; Park, K. K.; Park, Y. J.; Kim, W. H

    1999-08-01

    This report is intended to describe the statistical methods necessary to design and conduct radiation counting experiments and evaluate the data from the experiment. The methods are described for the evaluation of the stability of a counting system and the estimation of the precision of counting data by application of probability distribution models. The methods for the determination of the uncertainty of the results calculated from the number of counts, as well as various statistical methods for the reduction of counting error are also described. (Author). 11 refs., 8 tabs., 8 figs.

  20. Radiation counting statistics

    Energy Technology Data Exchange (ETDEWEB)

    Suh, M. Y.; Jee, K. Y.; Park, K. K. [Korea Atomic Energy Research Institute, Taejon (Korea)

    1999-08-01

    This report is intended to describe the statistical methods necessary to design and conduct radiation counting experiments and evaluate the data from the experiments. The methods are described for the evaluation of the stability of a counting system and the estimation of the precision of counting data by application of probability distribution models. The methods for the determination of the uncertainty of the results calculated from the number of counts, as well as various statistical methods for the reduction of counting error are also described. 11 refs., 6 figs., 8 tabs. (Author)

  1. Radiation counting statistics

    International Nuclear Information System (INIS)

    Suh, M. Y.; Jee, K. Y.; Park, K. K.; Park, Y. J.; Kim, W. H.

    1999-08-01

    This report is intended to describe the statistical methods necessary to design and conduct radiation counting experiments and evaluate the data from the experiment. The methods are described for the evaluation of the stability of a counting system and the estimation of the precision of counting data by application of probability distribution models. The methods for the determination of the uncertainty of the results calculated from the number of counts, as well as various statistical methods for the reduction of counting error are also described. (Author). 11 refs., 8 tabs., 8 figs

  2. Oil flow rate measurements using 198Au and total count technique

    International Nuclear Information System (INIS)

    Goncalves, Eduardo R.; Crispim, Verginia R.

    2013-01-01

    In industrial plants, oil and oil compounds are usually transported by closed pipelines with circular cross-section. The use of radiotracers in oil transport and processing industrial facilities allows calibrating flowmeters, measuring mean residence time in cracking columns, locate points of obstruction or leak in underground ducts, as well as investigating flow behavior or industrial processes such as in distillation towers. Inspection techniques using radiotracers are non-destructive, simple, economic and highly accurate. Among them, Total Count, which uses a small amount of radiotracer with known activity, is acknowledged as an absolute technique for flow rate measurement. A viscous fluid transport system, composed by four PVC pipelines with 13m length (12m horizontal and 1m vertical) and 1/2, 3/4, 1 and 2-inch gauges, respectively, interconnected by maneuvering valves was designed and assembled in order to conduct the research. This system was used to simulate different flow conditions of petroleum compounds and for experimental studies of flow profile in the horizontal and upward directions. As 198 Au presents a single photopeak (411,8 keV), it was the radioisotope chosen for oil labeling, in small amounts (6 ml) or around 200 kBq activity, and it was injected in the oil transport lines. A NaI scintillation detector 2'x 2', with well-defined geometry, was used to measure total activity, determine the calibration factor F and, positioned after a homogenization distance and interconnected to a standardized electronic set of nuclear instrumentation modules (NIM), to detect the radioactive cloud. (author)

  3. Limit of sensitivity of low-background counting equipment

    International Nuclear Information System (INIS)

    Homann, S.G.

    1991-01-01

    The Hazards Control Department's Radiological Measurements Laboratory (RML) analyzes many types of sample media in support of the Laboratory's health and safety program. The Department has determined that the equation for the minimum limit of sensitivity, MDC(α,β) = 2.71 + 3.29 (r b t s ) 1/2 is also adequate for RML counting systems with very-low-background levels. This paper reviews the normal distribution case and address the special case of determining the limit of sensitivity of a counting system when the background count rate is well known and small. In the latter case, we must use an exact test procedure based on the binomial distribution. However, the error in using the normal distribution for calculating a detection system's limit of sensitivity is not significant even as the total observed number of counts approaches or equals zero. 2 refs., 4 figs

  4. The Money/Counting Kit. The Prospectus Series, Paper No. 6.

    Science.gov (United States)

    Musumeci, Judith

    The Money/Counting Kit for Handicapped Children and Youth, frees the teacher from lessons in money and counting concepts and enables a student to learn at his own rate with immediate feedback from activity cards, name cards, thermoformed coin cards (optional), and self-instructional booklets. The activity cards, which may be used individually or…

  5. Relationship of Total Motile Sperm Count and Percentage Motile Sperm to Successful Pregnancy Rates Following Intrauterine Insemination

    OpenAIRE

    Pasqualotto, Eleonora B.; Daitch, James A.; Hendin, Benjamin N.; Falcone, Tommaso; Thomas, Anthony J.; Nelson, David R.; Agarwal, Ashok

    1999-01-01

    Purpose:This study sought (i) to investigate the relationship between postwash total motile sperm count and postwash percentage motile sperm in predicting successful intrauterine insemination and (ii) to determine the minimal postwash total motile sperm count required to achieve pregnancy with intrauterine insemination.

  6. The power and robustness of maximum LOD score statistics.

    Science.gov (United States)

    Yoo, Y J; Mendell, N R

    2008-07-01

    The maximum LOD score statistic is extremely powerful for gene mapping when calculated using the correct genetic parameter value. When the mode of genetic transmission is unknown, the maximum of the LOD scores obtained using several genetic parameter values is reported. This latter statistic requires higher critical value than the maximum LOD score statistic calculated from a single genetic parameter value. In this paper, we compare the power of maximum LOD scores based on three fixed sets of genetic parameter values with the power of the LOD score obtained after maximizing over the entire range of genetic parameter values. We simulate family data under nine generating models. For generating models with non-zero phenocopy rates, LOD scores maximized over the entire range of genetic parameters yielded greater power than maximum LOD scores for fixed sets of parameter values with zero phenocopy rates. No maximum LOD score was consistently more powerful than the others for generating models with a zero phenocopy rate. The power loss of the LOD score maximized over the entire range of genetic parameters, relative to the maximum LOD score calculated using the correct genetic parameter value, appeared to be robust to the generating models.

  7. Multiplicity counting from fission detector signals with time delay effects

    Science.gov (United States)

    Nagy, L.; Pázsit, I.; Pál, L.

    2018-03-01

    In recent work, we have developed the theory of using the first three auto- and joint central moments of the currents of up to three fission chambers to extract the singles, doubles and triples count rates of traditional multiplicity counting (Pázsit and Pál, 2016; Pázsit et al., 2016). The objective is to elaborate a method for determining the fissile mass, neutron multiplication, and (α, n) neutron emission rate of an unknown assembly of fissile material from the statistics of the fission chamber signals, analogous to the traditional multiplicity counting methods with detectors in the pulse mode. Such a method would be an alternative to He-3 detector systems, which would be free from the dead time problems that would be encountered in high counting rate applications, for example the assay of spent nuclear fuel. A significant restriction of our previous work was that all neutrons born in a source event (spontaneous fission) were assumed to be detected simultaneously, which is not fulfilled in reality. In the present work, this restriction is eliminated, by assuming an independent, identically distributed random time delay for all neutrons arising from one source event. Expressions are derived for the same auto- and joint central moments of the detector current(s) as in the previous case, expressed with the singles, doubles, and triples (S, D and T) count rates. It is shown that if the time-dispersion of neutron detections is of the same order of magnitude as the detector pulse width, as they typically are in measurements of fast neutrons, the multiplicity rates can still be extracted from the moments of the detector current, although with more involved calibration factors. The presented formulae, and hence also the performance of the proposed method, are tested by both analytical models of the time delay as well as with numerical simulations. Methods are suggested also for the modification of the method for large time delay effects (for thermalised neutrons).

  8. Simultaneous maximum a posteriori longitudinal PET image reconstruction

    Science.gov (United States)

    Ellis, Sam; Reader, Andrew J.

    2017-09-01

    Positron emission tomography (PET) is frequently used to monitor functional changes that occur over extended time scales, for example in longitudinal oncology PET protocols that include routine clinical follow-up scans to assess the efficacy of a course of treatment. In these contexts PET datasets are currently reconstructed into images using single-dataset reconstruction methods. Inspired by recently proposed joint PET-MR reconstruction methods, we propose to reconstruct longitudinal datasets simultaneously by using a joint penalty term in order to exploit the high degree of similarity between longitudinal images. We achieved this by penalising voxel-wise differences between pairs of longitudinal PET images in a one-step-late maximum a posteriori (MAP) fashion, resulting in the MAP simultaneous longitudinal reconstruction (SLR) method. The proposed method reduced reconstruction errors and visually improved images relative to standard maximum likelihood expectation-maximisation (ML-EM) in simulated 2D longitudinal brain tumour scans. In reconstructions of split real 3D data with inserted simulated tumours, noise across images reconstructed with MAP-SLR was reduced to levels equivalent to doubling the number of detected counts when using ML-EM. Furthermore, quantification of tumour activities was largely preserved over a variety of longitudinal tumour changes, including changes in size and activity, with larger changes inducing larger biases relative to standard ML-EM reconstructions. Similar improvements were observed for a range of counts levels, demonstrating the robustness of the method when used with a single penalty strength. The results suggest that longitudinal regularisation is a simple but effective method of improving reconstructed PET images without using resolution degrading priors.

  9. Effects of adipose tissue distribution on maximum lipid oxidation rate during exercise in normal-weight women.

    Science.gov (United States)

    Isacco, L; Thivel, D; Duclos, M; Aucouturier, J; Boisseau, N

    2014-06-01

    Fat mass localization affects lipid metabolism differently at rest and during exercise in overweight and normal-weight subjects. The aim of this study was to investigate the impact of a low vs high ratio of abdominal to lower-body fat mass (index of adipose tissue distribution) on the exercise intensity (Lipox(max)) that elicits the maximum lipid oxidation rate in normal-weight women. Twenty-one normal-weight women (22.0 ± 0.6 years, 22.3 ± 0.1 kg.m(-2)) were separated into two groups of either a low or high abdominal to lower-body fat mass ratio [L-A/LB (n = 11) or H-A/LB (n = 10), respectively]. Lipox(max) and maximum lipid oxidation rate (MLOR) were determined during a submaximum incremental exercise test. Abdominal and lower-body fat mass were determined from DXA scans. The two groups did not differ in aerobic fitness, total fat mass, or total and localized fat-free mass. Lipox(max) and MLOR were significantly lower in H-A/LB vs L-A/LB women (43 ± 3% VO(2max) vs 54 ± 4% VO(2max), and 4.8 ± 0.6 mg min(-1)kg FFM(-1)vs 8.4 ± 0.9 mg min(-1)kg FFM(-1), respectively; P normal-weight women, a predominantly abdominal fat mass distribution compared with a predominantly peripheral fat mass distribution is associated with a lower capacity to maximize lipid oxidation during exercise, as evidenced by their lower Lipox(max) and MLOR. Copyright © 2014 Elsevier Masson SAS. All rights reserved.

  10. Performance of in-pixel circuits for photon counting arrays (PCAs) based on polycrystalline silicon TFTs

    International Nuclear Information System (INIS)

    Liang, Albert K; Koniczek, Martin; Antonuk, Larry E; El-Mohri, Youcef; Zhao, Qihua; Street, Robert A; Lu, Jeng Ping

    2016-01-01

    Photon counting arrays (PCAs), defined as pixelated imagers which measure the absorbed energy of x-ray photons individually and record this information digitally, are of increasing clinical interest. A number of PCA prototypes with a 1 mm pixel-to-pixel pitch have recently been fabricated with polycrystalline silicon (poly-Si)—a thin-film technology capable of creating monolithic imagers of a size commensurate with human anatomy. In this study, analog and digital simulation frameworks were developed to provide insight into the influence of individual poly-Si transistors on pixel circuit performance—information that is not readily available through empirical means. The simulation frameworks were used to characterize the circuit designs employed in the prototypes. The analog framework, which determines the noise produced by individual transistors, was used to estimate energy resolution, as well as to identify which transistors contribute the most noise. The digital framework, which analyzes how well circuits function in the presence of significant variations in transistor properties, was used to estimate how fast a circuit can produce an output (referred to as output count rate). In addition, an algorithm was developed and used to estimate the minimum pixel pitch that could be achieved for the pixel circuits of the current prototypes. The simulation frameworks predict that the analog component of the PCA prototypes could have energy resolution as low as 8.9% full width at half maximum (FWHM) at 70 keV; and the digital components should work well even in the presence of significant thin-film transistor (TFT) variations, with the fastest component having output count rates as high as 3 MHz. Finally, based on conceivable improvements in the underlying fabrication process, the algorithm predicts that the 1 mm pitch of the current PCA prototypes could be reduced significantly, potentially to between ∼240 and 290 μm. (paper)

  11. Fast pulse discriminator for photon counting at high photon densities

    International Nuclear Information System (INIS)

    Benoit, R.; Pedrini, A.

    1977-03-01

    A fast tunnel diode discriminator for photon counting up to 200MHz count frequency is described. The tunnel diode is operated on its apparent I.V. characteristics displayed when the diode is driven into its oscillating region. The pulse shaper-discriminator is completely D.C. coupled in order to avoid base-line shift at high pulse rates

  12. Automated high performance liquid chromatography and liquid scintillation counting determination of pesticide mixture octanol/water partition rates

    International Nuclear Information System (INIS)

    Moody, R.P.; Carroll, J.M.; Kresta, A.M.

    1987-01-01

    Two novel methods are reported for measuring octanol/water partition rates of pesticides. A liquid scintillation counting (LSC) method was developed for automated monitoring of 14 C-labeled pesticides partitioning in biphasic water/octanol cocktail systems with limited success. A high performance liquid chromatography (HPLC) method was developed for automated partition rate monitoring of several constituents in a pesticide mixture, simultaneously. The mean log Kow +/- SD determined from triplicate experimental runs were for: 2,4-D-DMA (2,4-dichlorophenoxyacetic acid dimethylamine), 0.65 +/- .17; Deet (N,N-diethyl-m-toluamide), 2.02 +/- .01; Guthion (O,O-dimethyl-S-(4-oxo-1,2,3-benzotriazin-3(4H)-ylmethyl) phosphorodithioate), 2.43 +/- .03; Methyl-Parathion (O,O-dimethyl-O-(p-nitrophenyl) phosphorothioate), 2.68 +/- .05; and Fenitrothion (O,O-dimethyl O-(4-nitro-m-tolyl) phosphorothioate), 3.16 +/- .03. A strong positive linear correlation (r = .9979) was obtained between log Kow and log k' (log Kow = 2.35 (log k') + 0.63). The advantages that this automated procedure has in comparison with the standard manual shake-flask procedure are discussed

  13. Low-priced, time-saving, reliable and stable LR-115 counting system

    International Nuclear Information System (INIS)

    Tchorz-Trzeciakiewicz, D.E.

    2015-01-01

    Nuclear alpha particles leave etches (tracks) when they hit the surface of a LR-115 detector. The density of these tracks is used to measure radon concentration. Counting these tracks by human sense is tedious and time-consuming procedure and may introduce counting error, whereas most available automatic and semiautomatic counting systems are expensive or complex. An uncomplicated, robust, reliable and stable counting system using freely available on the Internet software as Digimizer™ and PhotoScape was developed and proposed. The effectiveness of the proposed procedure was evaluated by comparing the amount of tracks counted by software with the amount of tracks counted manually for 223 detectors. The percentage error for each analysed detector was obtained as a difference between automatic and manual counts divided by manual count. For more than 97% of detectors, the percentage errors oscillated between −3% and 3%. - Highlights: • Semiautomatic, uncomplicated procedure was proposed to count the amount of alpha tracks. • Freely available software on the Internet used as alpha tracks counting system for LR-115. • LR-115 detectors used to measure radon concentration and radon exhalation rate

  14. Determination of phosphorus-32 in wet-digested plant leaves by Cerenkov counting

    International Nuclear Information System (INIS)

    Wahid, P.A.; Kamalam, N.V.; Sankar, S.J.

    1985-01-01

    A method of determination of 32 P activity in leaf samples by Cerenkov counting technique is described. The method involves wet digestion of oven-dried leaves with 1:1 nitric-perchloric acid mixture followed by transferring the digest into a scintillation counting vial with distilled water upto a final volume of 20 mL, and determining the activity in a liquid scintillation system. Reproducible count rates can be obtained if the vials are counted after 4h allowing the silica in the digest to settle. (author)

  15. IDENTIFICATION OF BACTERIA CAUSING DIARRHOEA IN HIV/AIDS PATIENTS AND ITS CORRELATION WITH CD4 COUNT

    Directory of Open Access Journals (Sweden)

    Anand Premanand

    2016-05-01

    Full Text Available BACKGROUND The number of HIV-positive patients is increasing in India. Data on the prevalence of diarrhoea and the spectrum of bacteria responsible for diarrhoea in HIV- positive patients is lacking in our area. The identification of enteric pathogens in patients with HIV/AIDS is important because an increasing array of therapeutic regimens is becoming available to treat many of these infections. Thus, an attempt is done to elucidate the associations between causative bacteria of acute and chronic diarrhoea and CD4 count. METHODS Stool specimens were obtained over a period of eighteen months from HIV infected adults with diarrhoea presenting to Shri B M Patil Medical College Hospital and Research Centre, Vijayapura. In all patients with diarrhoea, stool specimens were examined by microscopy and cultures to identify bacterial pathogens and blood sample was analysed for CD4 count. RESULTS A total of 80 individuals were enrolled in this study. Cases included 46 males and 34 females. Among the cases, maximum subjects were found to be in the age group of 30-40 years in which 23 (62.2% were males and 14 (37.8% were females. 56 had acute and 24 had chronic diarrhoea. The percentages of bacteria isolated were 5 (8.9% in acute and 16 (66.7% in chronic diarrhoea respectively. The most common bacteria isolated was E. Coli (17.5% followed by Klebsiella (5% and Shigella Sps (3.75%. Patients with chronic diarrhoea had lower CD4 cell counts. The maximum bacterial isolation was in the patients whose CD4 cell counts were below 200 cells/mm3. CONCLUSION Bacterial isolation was most strongly associated with low CD4 counts and chronic diarrhoea. E. coli was isolated maximum among all the bacteria in the HIV patients. Over two-thirds of diarrhoeal episodes were undiagnosed, suggesting that unidentified agents or primary HIV enteropathy are important causes of diarrhoea in this population. There is a strong negative association between duration of diarrhoea and CD4

  16. Cerenkov counting and Cerenkov-scintillation counting with high refractive index organic liquids using a liquid scintillation counter

    Energy Technology Data Exchange (ETDEWEB)

    Wiebe, L I; Helus, F; Maier-Borst, W [Deutsches Krebsforschungszentrum, Heidelberg (Germany, F.R.). Inst. fuer Nuklearmedizin

    1978-06-01

    /sup 18/F and /sup 14/C radioactivity was measured in methyl salicylate (MS), a high refractive index hybrid Cherenkov-scintillation generating medium, using a liquid scintillation counter. At concentrations of up to 21.4%, in MS, dimethyl sulfoxide (DMSO) quenched /sup 14/C fluorescence, and with a 10-fold excess of DMSO over MS, /sup 18/F count rates were reduced below that for DMSO alone, probably as a result of concentration-independent self-quenching due to 'dark-complex' formation. DMSO in lower concentrations did not reduce the counting efficiency of /sup 18/F in MS. Nitrobenzene was a concentration-dependent quencher for both /sup 14/C and /sup 18/F in MS. Chlorobenzene (CB) and DMSO were both found to be weak Cherenkov generators with /sup 18/F. Counting efficiencies for /sup 18/F in MS, CB, and DMSO were 50.3, 7.8 and 4.3% respectively in the coincidence counting mode, and 58.1, 13.0 and 6.8% in the singles mode. /sup 14/C efficiencies were 14.4 and 22.3% for coincidence and singles respectively, and 15.3 and 42.0% using a modern counter designed for coincidence and single photon counting. The high /sup 14/C and /sup 18/F counting efficiency in MS are discussed with respect to excitation mechanism, on the basis of quench and channels ratios changes observed. It is proposed that MS functions as an efficient Cherenkov-scintillation generator for high-energy beta emitters such as /sup 18/F, and as a low-efficiency scintillator for weak beta emitting radionuclides such as /sup 14/C.

  17. Deep 3 GHz number counts from a P(D) fluctuation analysis

    Science.gov (United States)

    Vernstrom, T.; Scott, Douglas; Wall, J. V.; Condon, J. J.; Cotton, W. D.; Fomalont, E. B.; Kellermann, K. I.; Miller, N.; Perley, R. A.

    2014-05-01

    Radio source counts constrain galaxy populations and evolution, as well as the global star formation history. However, there is considerable disagreement among the published 1.4-GHz source counts below 100 μJy. Here, we present a statistical method for estimating the μJy and even sub-μJy source count using new deep wide-band 3-GHz data in the Lockman Hole from the Karl G. Jansky Very Large Array. We analysed the confusion amplitude distribution P(D), which provides a fresh approach in the form of a more robust model, with a comprehensive error analysis. We tested this method on a large-scale simulation, incorporating clustering and finite source sizes. We discuss in detail our statistical methods for fitting using Markov chain Monte Carlo, handling correlations, and systematic errors from the use of wide-band radio interferometric data. We demonstrated that the source count can be constrained down to 50 nJy, a factor of 20 below the rms confusion. We found the differential source count near 10 μJy to have a slope of -1.7, decreasing to about -1.4 at fainter flux densities. At 3 GHz, the rms confusion in an 8-arcsec full width at half-maximum beam is ˜ 1.2 μJy beam-1, and a radio background temperature ˜14 mK. Our counts are broadly consistent with published evolutionary models. With these results, we were also able to constrain the peak of the Euclidean normalized differential source count of any possible new radio populations that would contribute to the cosmic radio background down to 50 nJy.

  18. Correct liquid scintillation counting of steroids and glycosides in RIA samples: a comparison of xylene-based, dioxane-based and colloidal counting systems. Chapter 14

    International Nuclear Information System (INIS)

    Spolders, H.

    1977-01-01

    In RIA, the following parameters are important for accurate liquid scintillation counting. (1) Absence of chemiluminescence. (2) Stability of count rate. (3) Dissolving properties for the sample. For samples with varying colours, a quench correction must be applied. For any type of accurate quench correction, a homogeneous sample is necessary. This can be obtained if proteins and the buffer can be dissolved completely in the scintillator solution. In this paper, these criteria are compared in xylene-based, dioxane-based and colloidal scintillation solutions for either bound or free antigens of different polarity. The labelling radioisotope used was 3 H. Using colloidal scintillators with plasma and buffer samples, phasing or sedimentation of salt or proteins sometimes occurs. The influence of sedimentation or phasing on count rate stability and correct quench correction is illustrated by varying the ratio between the scintillator solution and a RIA sample containing a semi-polar steroid aldosterone. (author)

  19. Counting carbohydrates

    Science.gov (United States)

    Carb counting; Carbohydrate-controlled diet; Diabetic diet; Diabetes-counting carbohydrates ... Many foods contain carbohydrates (carbs), including: Fruit and fruit juice Cereal, bread, pasta, and rice Milk and milk products, soy milk Beans, legumes, ...

  20. How fast can quantum annealers count?

    International Nuclear Information System (INIS)

    Hen, Itay

    2014-01-01

    We outline an algorithm for the quantum counting problem using adiabatic quantum computation (AQC). We show that the mechanism of quantum-adiabatic evolution may be utilized toward estimating the number of solutions to a problem, and not only to find them. Using local adiabatic evolution, a process in which the adiabatic procedure is performed at a variable rate, the problem of counting the number of marked items in an unstructured database is solved quadratically faster than the corresponding classical algorithm. The above algorithm provides further evidence for the potentially powerful capabilities of AQC as a paradigm for more efficient problem solving on a quantum computer, and may be used as the basis for solving more sophisticated problems. (paper)

  1. Repeatability of differential goat bulk milk culture and associations with somatic cell count, total bacterial count, and standard plate count

    OpenAIRE

    Koop, G.; Dik, N.; Nielen, M.; Lipman, L.J.A.

    2010-01-01

    The aims of this study were to assess how different bacterial groups in bulk milk are related to bulk milk somatic cell count (SCC), bulk milk total bacterial count (TBC), and bulk milk standard plate count (SPC) and to measure the repeatability of bulk milk culturing. On 53 Dutch dairy goat farms, 3 bulk milk samples were collected at intervals of 2 wk. The samples were cultured for SPC, coliform count, and staphylococcal count and for the presence of Staphylococcus aureus. Furthermore, SCC ...

  2. Clustering method for counting passengers getting in a bus with single camera

    Science.gov (United States)

    Yang, Tao; Zhang, Yanning; Shao, Dapei; Li, Ying

    2010-03-01

    Automatic counting of passengers is very important for both business and security applications. We present a single-camera-based vision system that is able to count passengers in a highly crowded situation at the entrance of a traffic bus. The unique characteristics of the proposed system include, First, a novel feature-point-tracking- and online clustering-based passenger counting framework, which performs much better than those of background-modeling-and foreground-blob-tracking-based methods. Second, a simple and highly accurate clustering algorithm is developed that projects the high-dimensional feature point trajectories into a 2-D feature space by their appearance and disappearance times and counts the number of people through online clustering. Finally, all test video sequences in the experiment are captured from a real traffic bus in Shanghai, China. The results show that the system can process two 320×240 video sequences at a frame rate of 25 fps simultaneously, and can count passengers reliably in various difficult scenarios with complex interaction and occlusion among people. The method achieves high accuracy rates up to 96.5%.

  3. Radon Daughters Background Reduction in Alpha Particles Counting System

    International Nuclear Information System (INIS)

    Dadon, S. S.; Pelled, O.; Orion, I.

    2014-01-01

    The ABPC method is using a serially occurring events of the beta decay of the 214Bi fallow by alpha decay of the 214Po that take place almost simultaneously to detect the Pseudo Coincidence Event (PCE) from the RDP, and to subtract them from the gross alpha counts. 267 This work showed that it is possible to improve the efficiency of RDP background reduction, including subtracting the 218Po contribution by using the ABPC method based on a single solid state silicon PIPS detector. False counts percentage obtained at the output of the PCE circuit were smaller than 0.1%. The results show that the PCE circuit was not influenced by non RDP alpha emitters. The PCE system did not reduce the non PCE of the 218Po. After 20 minutes the 218Po was strongly decayed, and its contribution became negligible. In order to overcome this disadvantage, a mathematical matching calculations for the 214Po and the 218Po decay equations were employed, and a constant ratio of the APo214(0) / APo218(0) was obtained. This ratio can be used to estimate the count rate of the 218Po at the first 20 minutes, and to subtract it from the total count rate in order to obtain correct RDP reduction

  4. Calibration of a liquid scintillation counter for alpha, beta and Cerenkov counting

    International Nuclear Information System (INIS)

    Scarpitta, S.C.; Fisenne, I.M.

    1996-07-01

    Calibration data are presented for 25 radionuclides that were individually measured in a Packard Tri-Carb 2250CA liquid scintillation (LS) counter by both conventional and Cerenkov detection techniques. The relationships and regression data between the quench indicating parameters and the LS counting efficiencies were determined using microliter amounts of tracer added to low 40 K borosilicate glass vials containing 15 mL of Insta-Gel XF scintillation cocktail. Using 40 K, the detection efficiencies were linear over a three order of magnitude range (10 - 10,000 mBq) in beta activity for both LS and Cerenkov counting. The Cerenkov counting efficiency (CCE) increased linearly (42% per MeV) from 0.30 to 2.0 MeV, whereas the LS efficiency was >90% for betas with energy in excess of 0.30 MeV. The CCE was 20 - 50% less than the LS counting efficiency for beta particles with maximum energies in excess of 1 MeV. Based on replicate background measurements, the lower limit of detection (LLD) for a 1-h count at the 95% confidence level, using water as a solvent, was 0.024 counts sec- -1 and 0.028 counts sec-1 for plastic and glass vials, respectively. The LLD for a 1-h-count ranged from 46 to 56 mBq (2.8 - 3.4 dpm) for both Cerenkov and conventional LS counting. This assumes: (1) a 100% counting efficiency, (2) a 50% yield of the nuclide of interest, (3) a 1-h measurement time using low background plastic vials, and (4) a 0-50 keV region of interest. The LLD is reduced an order of magnitude when the yield recovery exceeds 90% and a lower background region is used (i.e., 100 - 500 keV alpha region of interest). Examples and applications of both Cerenkov and LS counting techniques are given in the text and appendices

  5. Scale-dependent associations of Band-tailed Pigeon counts at mineral sites

    Science.gov (United States)

    Overton, Cory T.; Casazza, Michael L.; Coates, Peter S.

    2010-01-01

    The abundance of Band-tailed Pigeons (Patagioenas fasciata monilis) has declined substantially from historic numbers along the Pacific Coast. Identification of patterns and causative factors of this decline are hampered because habitat use data are limited, and temporal and spatial variability patterns associated with population indices are not known. Furthermore, counts are influenced not only by pigeon abundance but also by rate of visitation to mineral sites, which may not be consistent. To address these issues, we conducted mineral site counts during 2001 and 2002 at 20 locations from 4 regions in the Pacific Northwest, including central Oregon and western Washington, USA, and British Columbia, Canada. We developed inference models that consisted of environmental factors and spatial characteristics at multiple spatial scales. Based on information theory, we compared models within a final set that included variables measured at 3 spatial scales (0.03 ha, 3.14 ha, and 7850 ha). Pigeon counts increased from central Oregon through northern Oregon and decreased into British Columbia. After accounting for this spatial pattern, we found that pigeon counts increased 12% ± 2.7 with a 10% increase in the amount of deciduous forested area within 100 m from a mineral site. Also, distance from the mineral site of interest to the nearest known mineral site was positively related to pigeon counts. These findings provide direction for future research focusing on understanding the relationships between indices of relative abundance and complete counts (censuses) of pigeon populations by identifying habitat characteristics that might influence visitation rates. Furthermore, our results suggest that spatial arrangement of mineral sites influences Band-tailed Pigeon counts and the populations which those counts represent.

  6. Development of a phoswich detector for neutron dose rate measurements in the Earth's atmosphere

    International Nuclear Information System (INIS)

    Doensdorf, Esther Miriam

    2014-01-01

    dose rates is determined. The count rates show a maximum which corresponds to the location of the Pfotzer maximum (Pfotzer, 1936). In the dose rates a maximum is found as well, which has been mentioned in Bartlett (2004).

  7. Comparative Effectiveness of Two Walking Interventions on Participation, Step Counts, and Health.

    Science.gov (United States)

    Smith-McLallen, Aaron; Heller, Debbie; Vernisi, Kristin; Gulick, Diana; Cruz, Samantha; Snyder, Richard L

    2017-03-01

    To (1) compare the effects of two worksite-based walking interventions on employee participation rates; (2) compare average daily step counts between conditions, and; (3) examine the effects of increases in average daily step counts on biometric and psychologic outcomes. We conducted a cluster-randomized trial in which six employer groups were randomly selected and randomly assigned to condition. Four manufacturing worksites and two office-based worksite served as the setting. A total of 474 employees from six employer groups were included. A standard walking program was compared to an enhanced program that included incentives, feedback, competitive challenges, and monthly wellness workshops. Walking was measured by self-reported daily step counts. Survey measures and biometric screenings were administered at baseline and 3, 6, and 9 months after baseline. Analysis used linear mixed models with repeated measures. During 9 months, participants in the enhanced condition averaged 726 more steps per day compared with those in the standard condition (p women (-2.1 lbs.), and reductions in body mass index (-0.41 men, -0.31 women). Higher step counts were also associated with improvements in mood, having more energy, and higher ratings of overall health. An enhanced walking program significantly increases participation rates and daily step counts, which were associated with weight loss and reductions in body mass index.

  8. Enhanced coulomb counting method for estimating state-of-charge and state-of-health of lithium-ion batteries

    International Nuclear Information System (INIS)

    Ng, Kong Soon; Moo, Chin-Sien; Chen, Yi-Ping; Hsieh, Yao-Ching

    2009-01-01

    The coulomb counting method is expedient for state-of-charge (SOC) estimation of lithium-ion batteries with high charging and discharging efficiencies. The charging and discharging characteristics are investigated and reveal that the coulomb counting method is convenient and accurate for estimating the SOC of lithium-ion batteries. A smart estimation method based on coulomb counting is proposed to improve the estimation accuracy. The corrections are made by considering the charging and operating efficiencies. Furthermore, the state-of-health (SOH) is evaluated by the maximum releasable capacity. Through the experiments that emulate practical operations, the SOC estimation method is verified to demonstrate the effectiveness and accuracy.

  9. Alabama Kids Count 2002 Data Book.

    Science.gov (United States)

    Curtis, Apreill; Bogie, Don

    This Kids Count data book examines statewide trends in well-being of Alabamas children. The statistical portrait is based on 18 indicators in the areas of child health, education, safety, and security: (1) infant mortality rate; (2) low weight births; (3) child health index; (4) births to unmarried teens; (5) first grade retention; (6) school…

  10. Regression analysis of mixed recurrent-event and panel-count data.

    Science.gov (United States)

    Zhu, Liang; Tong, Xinwei; Sun, Jianguo; Chen, Manhua; Srivastava, Deo Kumar; Leisenring, Wendy; Robison, Leslie L

    2014-07-01

    In event history studies concerning recurrent events, two types of data have been extensively discussed. One is recurrent-event data (Cook and Lawless, 2007. The Analysis of Recurrent Event Data. New York: Springer), and the other is panel-count data (Zhao and others, 2010. Nonparametric inference based on panel-count data. Test 20: , 1-42). In the former case, all study subjects are monitored continuously; thus, complete information is available for the underlying recurrent-event processes of interest. In the latter case, study subjects are monitored periodically; thus, only incomplete information is available for the processes of interest. In reality, however, a third type of data could occur in which some study subjects are monitored continuously, but others are monitored periodically. When this occurs, we have mixed recurrent-event and panel-count data. This paper discusses regression analysis of such mixed data and presents two estimation procedures for the problem. One is a maximum likelihood estimation procedure, and the other is an estimating equation procedure. The asymptotic properties of both resulting estimators of regression parameters are established. Also, the methods are applied to a set of mixed recurrent-event and panel-count data that arose from a Childhood Cancer Survivor Study and motivated this investigation. © The Author 2014. Published by Oxford University Press. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

  11. Maximum Entropy in Drug Discovery

    Directory of Open Access Journals (Sweden)

    Chih-Yuan Tseng

    2014-07-01

    Full Text Available Drug discovery applies multidisciplinary approaches either experimentally, computationally or both ways to identify lead compounds to treat various diseases. While conventional approaches have yielded many US Food and Drug Administration (FDA-approved drugs, researchers continue investigating and designing better approaches to increase the success rate in the discovery process. In this article, we provide an overview of the current strategies and point out where and how the method of maximum entropy has been introduced in this area. The maximum entropy principle has its root in thermodynamics, yet since Jaynes’ pioneering work in the 1950s, the maximum entropy principle has not only been used as a physics law, but also as a reasoning tool that allows us to process information in hand with the least bias. Its applicability in various disciplines has been abundantly demonstrated. We give several examples of applications of maximum entropy in different stages of drug discovery. Finally, we discuss a promising new direction in drug discovery that is likely to hinge on the ways of utilizing maximum entropy.

  12. Photon counting and fluctuation of molecular movement

    International Nuclear Information System (INIS)

    Inohara, Koichi

    1978-01-01

    The direct measurement of the fluctuation of molecular motions, which provides with useful information on the molecular movement, was conducted by introducing photon counting method. The utilization of photon counting makes it possible to treat the molecular system consisting of a small number of molecules like a radioisotope in the detection of a small number of atoms, which are significant in biological systems. This method is based on counting the number of photons of the definite polarization emitted in a definite time interval from the fluorescent molecules excited by pulsed light, which are bound to the marked large molecules found in a definite spatial region. Using the probability of finding a number of molecules oriented in a definite direction in the definite spatial region, the probability of counting a number of photons in a definite time interval can be calculated. Thus the measurable count rate of photons can be related with the fluctuation of molecular movement. The measurement was carried out under the condition, in which the probability of the simultaneous arrival of more than two photons at a detector is less than 1/100. As the experimental results, the resolving power of photon-counting apparatus, the frequency distribution of the number of photons of some definite polarization counted for 1 nanosecond are shown. In the solution, the variance of the number of molecules of 500 on the average is 1200, which was estimated from the experimental data by assuming normal distribution. This departure from the Poisson distribution means that a certain correlation does exist in molecular movement. In solid solution, no significant deviation was observed. The correlation existing in molecular movement can be expressed in terms of the fluctuation of the number of molecules. (Nakai, Y.)

  13. Analysis of overdispersed count data by mixtures of Poisson variables and Poisson processes.

    Science.gov (United States)

    Hougaard, P; Lee, M L; Whitmore, G A

    1997-12-01

    Count data often show overdispersion compared to the Poisson distribution. Overdispersion is typically modeled by a random effect for the mean, based on the gamma distribution, leading to the negative binomial distribution for the count. This paper considers a larger family of mixture distributions, including the inverse Gaussian mixture distribution. It is demonstrated that it gives a significantly better fit for a data set on the frequency of epileptic seizures. The same approach can be used to generate counting processes from Poisson processes, where the rate or the time is random. A random rate corresponds to variation between patients, whereas a random time corresponds to variation within patients.

  14. Standardization of 241Am by digital coincidence counting, liquid scintillation counting and defined solid angle counting

    International Nuclear Information System (INIS)

    Balpardo, C.; Capoulat, M.E.; Rodrigues, D.; Arenillas, P.

    2010-01-01

    The nuclide 241 Am decays by alpha emission to 237 Np. Most of the decays (84.6%) populate the excited level of 237 Np with energy of 59.54 keV. Digital coincidence counting was applied to standardize a solution of 241 Am by alpha-gamma coincidence counting with efficiency extrapolation. Electronic discrimination was implemented with a pressurized proportional counter and the results were compared with two other independent techniques: Liquid scintillation counting using the logical sum of double coincidences in a TDCR array and defined solid angle counting taking into account activity inhomogeneity in the active deposit. The results show consistency between the three methods within a limit of a 0.3%. An ampoule of this solution will be sent to the International Reference System (SIR) during 2009. Uncertainties were analysed and compared in detail for the three applied methods.

  15. Statistical Methods for Unusual Count Data

    DEFF Research Database (Denmark)

    Guthrie, Katherine A.; Gammill, Hilary S.; Kamper-Jørgensen, Mads

    2016-01-01

    microchimerism data present challenges for statistical analysis, including a skewed distribution, excess zero values, and occasional large values. Methods for comparing microchimerism levels across groups while controlling for covariates are not well established. We compared statistical models for quantitative...... microchimerism values, applied to simulated data sets and 2 observed data sets, to make recommendations for analytic practice. Modeling the level of quantitative microchimerism as a rate via Poisson or negative binomial model with the rate of detection defined as a count of microchimerism genome equivalents per...

  16. Evolution of high tooth replacement rates in sauropod dinosaurs.

    Science.gov (United States)

    D'Emic, Michael D; Whitlock, John A; Smith, Kathlyn M; Fisher, Daniel C; Wilson, Jeffrey A

    2013-01-01

    Tooth replacement rate can be calculated in extinct animals by counting incremental lines of deposition in tooth dentin. Calculating this rate in several taxa allows for the study of the evolution of tooth replacement rate. Sauropod dinosaurs, the largest terrestrial animals that ever evolved, exhibited a diversity of tooth sizes and shapes, but little is known about their tooth replacement rates. We present tooth replacement rate, formation time, crown volume, total dentition volume, and enamel thickness for two coexisting but distantly related and morphologically disparate sauropod dinosaurs Camarasaurus and Diplodocus. Individual tooth formation time was determined by counting daily incremental lines in dentin. Tooth replacement rate is calculated as the difference between the number of days recorded in successive replacement teeth. Each tooth family in Camarasaurus has a maximum of three replacement teeth, whereas each Diplodocus tooth family has up to five. Tooth formation times are about 1.7 times longer in Camarasaurus than in Diplodocus (315 vs. 185 days). Average tooth replacement rate in Camarasaurus is about one tooth every 62 days versus about one tooth every 35 days in Diplodocus. Despite slower tooth replacement rates in Camarasaurus, the volumetric rate of Camarasaurus tooth replacement is 10 times faster than in Diplodocus because of its substantially greater tooth volumes. A novel method to estimate replacement rate was developed and applied to several other sauropodomorphs that we were not able to thin section. Differences in tooth replacement rate among sauropodomorphs likely reflect disparate feeding strategies and/or food choices, which would have facilitated the coexistence of these gigantic herbivores in one ecosystem. Early neosauropods are characterized by high tooth replacement rates (despite their large tooth size), and derived titanosaurs and diplodocoids independently evolved the highest known tooth replacement rates among archosaurs.

  17. Evolution of high tooth replacement rates in sauropod dinosaurs.

    Directory of Open Access Journals (Sweden)

    Michael D D'Emic

    Full Text Available BACKGROUND: Tooth replacement rate can be calculated in extinct animals by counting incremental lines of deposition in tooth dentin. Calculating this rate in several taxa allows for the study of the evolution of tooth replacement rate. Sauropod dinosaurs, the largest terrestrial animals that ever evolved, exhibited a diversity of tooth sizes and shapes, but little is known about their tooth replacement rates. METHODOLOGY/PRINCIPAL FINDINGS: We present tooth replacement rate, formation time, crown volume, total dentition volume, and enamel thickness for two coexisting but distantly related and morphologically disparate sauropod dinosaurs Camarasaurus and Diplodocus. Individual tooth formation time was determined by counting daily incremental lines in dentin. Tooth replacement rate is calculated as the difference between the number of days recorded in successive replacement teeth. Each tooth family in Camarasaurus has a maximum of three replacement teeth, whereas each Diplodocus tooth family has up to five. Tooth formation times are about 1.7 times longer in Camarasaurus than in Diplodocus (315 vs. 185 days. Average tooth replacement rate in Camarasaurus is about one tooth every 62 days versus about one tooth every 35 days in Diplodocus. Despite slower tooth replacement rates in Camarasaurus, the volumetric rate of Camarasaurus tooth replacement is 10 times faster than in Diplodocus because of its substantially greater tooth volumes. A novel method to estimate replacement rate was developed and applied to several other sauropodomorphs that we were not able to thin section. CONCLUSIONS/SIGNIFICANCE: Differences in tooth replacement rate among sauropodomorphs likely reflect disparate feeding strategies and/or food choices, which would have facilitated the coexistence of these gigantic herbivores in one ecosystem. Early neosauropods are characterized by high tooth replacement rates (despite their large tooth size, and derived titanosaurs and

  18. Alabama Kids Count 2001 Data Book.

    Science.gov (United States)

    Curtis, Apreill; Bogie, Don

    This Kids Count data book examines statewide trends in well-being for Alabama's children. The statistical portrait is based on 17 indicators in the areas of health, education, safety, and security. The indicators are: (1) infant mortality rate; (2) low weight births; (3) child health index; (4) births to unmarried teens; (5) first grade retention;…

  19. Nevada Kids Count Data Book, 1997.

    Science.gov (United States)

    We Can, Inc., Las Vegas, NV.

    This Kids Count data book is the first to examine statewide indicators of the well being of Nevada's children. The statistical portrait is based on 15 indicators of child well being: (1) percent low birth-weight babies; (2) infant mortality rate; (3) percent of children in poverty; (4) percent of children in single-parent families; (5) percent of…

  20. Benjamin Thompson, Count Rumford Count Rumford on the nature of heat

    CERN Document Server

    Brown, Sanborn C

    1967-01-01

    Men of Physics: Benjamin Thompson - Count Rumford: Count Rumford on the Nature of Heat covers the significant contributions of Count Rumford in the fields of physics. Count Rumford was born with the name Benjamin Thompson on March 23, 1753, in Woburn, Massachusetts. This book is composed of two parts encompassing 11 chapters, and begins with a presentation of Benjamin Thompson's biography and his interest in physics, particularly as an advocate of an """"anti-caloric"""" theory of heat. The subsequent chapters are devoted to his many discoveries that profoundly affected the physical thought

  1. Evaluation of bias and variance in low-count OSEM list mode reconstruction

    International Nuclear Information System (INIS)

    Jian, Y; Carson, R E; Planeta, B

    2015-01-01

    Statistical algorithms have been widely used in PET image reconstruction. The maximum likelihood expectation maximization reconstruction has been shown to produce bias in applications where images are reconstructed from a relatively small number of counts. In this study, image bias and variability in low-count OSEM reconstruction are investigated on images reconstructed with MOLAR (motion-compensation OSEM list-mode algorithm for resolution-recovery reconstruction) platform. A human brain ([ 11 C]AFM) and a NEMA phantom are used in the simulation and real experiments respectively, for the HRRT and Biograph mCT. Image reconstructions were repeated with different combinations of subsets and iterations. Regions of interest were defined on low-activity and high-activity regions to evaluate the bias and noise at matched effective iteration numbers (iterations × subsets). Minimal negative biases and no positive biases were found at moderate count levels and less than 5% negative bias was found using extremely low levels of counts (0.2 M NEC). At any given count level, other factors, such as subset numbers and frame-based scatter correction may introduce small biases (1–5%) in the reconstructed images. The observed bias was substantially lower than that reported in the literature, perhaps due to the use of point spread function and/or other implementation methods in MOLAR. (paper)

  2. CLARO: an ASIC for high rate single photon counting with multi-anode photomultipliers

    Science.gov (United States)

    Baszczyk, M.; Carniti, P.; Cassina, L.; Cotta Ramusino, A.; Dorosz, P.; Fiorini, M.; Gotti, C.; Kucewicz, W.; Malaguti, R.; Pessina, G.

    2017-08-01

    The CLARO is a radiation-hard 8-channel ASIC designed for single photon counting with multi-anode photomultiplier tubes. Each channel outputs a digital pulse when the input signal from the photomultiplier crosses a configurable threshold. The fast return to baseline, typically within 25 ns, and below 50 ns in all conditions, allows to count up to 107 hits/s on each channel, with a power consumption of about 1 mW per channel. The ASIC presented here is a much improved version of the first 4-channel prototype. The threshold can be precisely set in a wide range, between 30 ke- (5 fC) and 16 Me- (2.6 pC). The noise of the amplifier with a 10 pF input capacitance is 3.5 ke- (0.6 fC) RMS. All settings are stored in a 128-bit configuration and status register, protected against soft errors with triple modular redundancy. The paper describes the design of the ASIC at transistor-level, and demonstrates its performance on the test bench.

  3. Inferior clinical outcome of the CD4+ cell count-guided antiretroviral treatment interruption strategy in the SMART study: role of CD4+ Cell counts and HIV RNA levels during follow-up

    DEFF Research Database (Denmark)

    Lundgren, Jens; Babiker, Abdel; El-Sadr, Wafaa

    2008-01-01

    BACKGROUND AND METHODS: The SMART study compared 2 strategies for using antiretroviral therapy-drug conservation (DC) and viral suppression (VS)-in 5,472 human immunodeficiency virus (HIV)-infected patients with CD4+ cell counts >350 cells/microL. Rates and predictors of opportunistic disease...... or death (OD/death) and the relative risk (RR) in DC versus VS groups according to the latest CD4+ cell count and HIV RNA level are reported. RESULTS: During a mean of 16 months of follow-up, DC patients spent more time with a latest CD4+ cell count ...%) and with a latest HIV RNA level >400 copies/mL (71% vs. 28%) and had a higher rate of OD/death (3.4 vs. 1.3/100 person-years) than VS patients. For periods of follow- up with a CD4+ cell count

  4. Patient-dependent count-rate adaptive normalization for PET detector efficiency with delayed-window coincidence events

    International Nuclear Information System (INIS)

    Niu, Xiaofeng; Ye, Hongwei; Xia, Ting; Asma, Evren; Gagnon, Daniel; Wang, Wenli; Winkler, Mark

    2015-01-01

    Quantitative PET imaging is widely used in clinical diagnosis in oncology and neuroimaging. Accurate normalization correction for the efficiency of each line-of- response is essential for accurate quantitative PET image reconstruction. In this paper, we propose a normalization calibration method by using the delayed-window coincidence events from the scanning phantom or patient. The proposed method could dramatically reduce the ‘ring’ artifacts caused by mismatched system count-rates between the calibration and phantom/patient datasets. Moreover, a modified algorithm for mean detector efficiency estimation is proposed, which could generate crystal efficiency maps with more uniform variance. Both phantom and real patient datasets are used for evaluation. The results show that the proposed method could lead to better uniformity in reconstructed images by removing ring artifacts, and more uniform axial variance profiles, especially around the axial edge slices of the scanner. The proposed method also has the potential benefit to simplify the normalization calibration procedure, since the calibration can be performed using the on-the-fly acquired delayed-window dataset. (paper)

  5. Daily step count predicts acute exacerbations in a US cohort with COPD.

    Science.gov (United States)

    Moy, Marilyn L; Teylan, Merilee; Weston, Nicole A; Gagnon, David R; Garshick, Eric

    2013-01-01

    COPD is characterized by variability in exercise capacity and physical activity (PA), and acute exacerbations (AEs). Little is known about the relationship between daily step count, a direct measure of PA, and the risk of AEs, including hospitalizations. In an observational cohort study of 169 persons with COPD, we directly assessed PA with the StepWatch Activity Monitor, an ankle-worn accelerometer that measures daily step count. We also assessed exercise capacity with the 6-minute walk test (6MWT) and patient-reported PA with the St. George's Respiratory Questionnaire Activity Score (SGRQ-AS). AEs and COPD-related hospitalizations were assessed and validated prospectively over a median of 16 months. Mean daily step count was 5804±3141 steps. Over 209 person-years of observation, there were 263 AEs (incidence rate 1.3±1.6 per person-year) and 116 COPD-related hospitalizations (incidence rate 0.56±1.09 per person-year). Adjusting for FEV1 % predicted and prednisone use for AE in previous year, for each 1000 fewer steps per day walked at baseline, there was an increased rate of AEs (rate ratio 1.07; 95%CI = 1.003-1.15) and COPD-related hospitalizations (rate ratio 1.24; 95%CI = 1.08-1.42). There was a significant linear trend of decreasing daily step count by quartiles and increasing rate ratios for AEs (P = 0.008) and COPD-related hospitalizations (P = 0.003). Each 30-meter decrease in 6MWT distance was associated with an increased rate ratio of 1.07 (95%CI = 1.01-1.14) for AEs and 1.18 (95%CI = 1.07-1.30) for COPD-related hospitalizations. Worsening of SGRQ-AS by 4 points was associated with an increased rate ratio of 1.05 (95%CI = 1.01-1.09) for AEs and 1.10 (95%CI = 1.02-1.17) for COPD-related hospitalizations. Lower daily step count, lower 6MWT distance, and worse SGRQ-AS predict future AEs and COPD-related hospitalizations, independent of pulmonary function and previous AE history. These results support the importance of

  6. Energy-correction photon counting pixel for photon energy extraction under pulse pile-up

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Daehee; Park, Kyungjin; Lim, Kyung Taek; Cho, Gyuseong, E-mail: gscho@kaist.ac.kr

    2017-06-01

    A photon counting detector (PCD) has been proposed as an alternative solution to an energy-integrating detector (EID) in medical imaging field due to its high resolution, high efficiency, and low noise. The PCD has expanded to variety of fields such as spectral CT, k-edge imaging, and material decomposition owing to its capability to count and measure the number and the energy of an incident photon, respectively. Nonetheless, pulse pile-up, which is a superimposition of pulses at the output of a charge sensitive amplifier (CSA) in each PC pixel, occurs frequently as the X-ray flux increases due to the finite pulse processing time (PPT) in CSAs. Pulse pile-up induces not only a count loss but also distortion in the measured X-ray spectrum from each PC pixel and thus it is a main constraint on the use of PCDs in high flux X-ray applications. To minimize these effects, an energy-correction PC (ECPC) pixel is proposed to resolve pulse pile-up without cutting off the PPT by adding an energy correction logic (ECL) via a cross detection method (CDM). The ECPC pixel with a size of 200×200 µm{sup 2} was fabricated by using a 6-metal 1-poly 0.18 µm CMOS process with a static power consumption of 7.2 μW/pixel. The maximum count rate of the ECPC pixel was extended by approximately three times higher than that of a conventional PC pixel with a PPT of 500 nsec. The X-ray spectrum of 90 kVp, filtered by 3 mm Al filter, was measured as the X-ray current was increased using the CdTe and the ECPC pixel. As a result, the ECPC pixel dramatically reduced the energy spectrum distortion at 2 Mphotons/pixel/s when compared to that of the ERCP pixel with the same 500 nsec PPT.

  7. Counting cormorants

    DEFF Research Database (Denmark)

    Bregnballe, Thomas; Carss, David N; Lorentsen, Svein-Håkon

    2013-01-01

    This chapter focuses on Cormorant population counts for both summer (i.e. breeding) and winter (i.e. migration, winter roosts) seasons. It also explains differences in the data collected from undertaking ‘day’ versus ‘roost’ counts, gives some definitions of the term ‘numbers’, and presents two...

  8. The mechanics of granitoid systems and maximum entropy production rates.

    Science.gov (United States)

    Hobbs, Bruce E; Ord, Alison

    2010-01-13

    A model for the formation of granitoid systems is developed involving melt production spatially below a rising isotherm that defines melt initiation. Production of the melt volumes necessary to form granitoid complexes within 10(4)-10(7) years demands control of the isotherm velocity by melt advection. This velocity is one control on the melt flux generated spatially just above the melt isotherm, which is the control valve for the behaviour of the complete granitoid system. Melt transport occurs in conduits initiated as sheets or tubes comprising melt inclusions arising from Gurson-Tvergaard constitutive behaviour. Such conduits appear as leucosomes parallel to lineations and foliations, and ductile and brittle dykes. The melt flux generated at the melt isotherm controls the position of the melt solidus isotherm and hence the physical height of the Transport/Emplacement Zone. A conduit width-selection process, driven by changes in melt viscosity and constitutive behaviour, operates within the Transport Zone to progressively increase the width of apertures upwards. Melt can also be driven horizontally by gradients in topography; these horizontal fluxes can be similar in magnitude to vertical fluxes. Fluxes induced by deformation can compete with both buoyancy and topographic-driven flow over all length scales and results locally in transient 'ponds' of melt. Pluton emplacement is controlled by the transition in constitutive behaviour of the melt/magma from elastic-viscous at high temperatures to elastic-plastic-viscous approaching the melt solidus enabling finite thickness plutons to develop. The system involves coupled feedback processes that grow at the expense of heat supplied to the system and compete with melt advection. The result is that limits are placed on the size and time scale of the system. Optimal characteristics of the system coincide with a state of maximum entropy production rate. This journal is © 2010 The Royal Society

  9. Modelling maximum canopy conductance and transpiration in ...

    African Journals Online (AJOL)

    There is much current interest in predicting the maximum amount of water that can be transpired by Eucalyptus trees. It is possible that industrial waste water may be applied as irrigation water to eucalypts and it is important to predict the maximum transpiration rates of these plantations in an attempt to dispose of this ...

  10. A simulation study of Linsley's approach to infer elongation rate and fluctuations of the EAS maximum depth from muon arrival time distributions

    International Nuclear Information System (INIS)

    Badea, A.F.; Brancus, I.M.; Rebel, H.; Haungs, A.; Oehlschlaeger, J.; Zazyan, M.

    1999-01-01

    The average depth of the maximum X m of the EAS (Extensive Air Shower) development depends on the energy E 0 and the mass of the primary particle, and its dependence from the energy is traditionally expressed by the so-called elongation rate D e defined as change in the average depth of the maximum per decade of E 0 i.e. D e = dX m /dlog 10 E 0 . Invoking the superposition model approximation i.e. assuming that a heavy primary (A) has the same shower elongation rate like a proton, but scaled with energies E 0 /A, one can write X m = X init + D e log 10 (E 0 /A). In 1977 an indirect approach studying D e has been suggested by Linsley. This approach can be applied to shower parameters which do not depend explicitly on the energy of the primary particle, but do depend on the depth of observation X and on the depth X m of shower maximum. The distribution of the EAS muon arrival times, measured at a certain observation level relatively to the arrival time of the shower core reflect the pathlength distribution of the muon travel from locus of production (near the axis) to the observation locus. The basic a priori assumption is that we can associate the mean value or median T of the time distribution to the height of the EAS maximum X m , and that we can express T = f(X,X m ). In order to derive from the energy variation of the arrival time quantities information about elongation rate, some knowledge is required about F i.e. F = - ∂ T/∂X m ) X /∂(T/∂X) X m , in addition to the variations with the depth of observation and the zenith-angle (θ) dependence, respectively. Thus ∂T/∂log 10 E 0 | X = - F·D e ·1/X v ·∂T/∂secθ| E 0 . In a similar way the fluctuations σ(X m ) of X m may be related to the fluctuations σ(T) of T i.e. σ(T) = - σ(X m )· F σ ·1/X v ·∂T/∂secθ| E 0 , with F σ being the corresponding scaling factor for the fluctuation of F. By simulations of the EAS development using the Monte Carlo code CORSIKA the energy and angle

  11. Current automated 3D cell detection methods are not a suitable replacement for manual stereologic cell counting

    Directory of Open Access Journals (Sweden)

    Christoph eSchmitz

    2014-05-01

    Full Text Available Stereologic cell counting has had a major impact on the field of neuroscience. A major bottleneck in stereologic cell counting is that the user must manually decide whether or not each cell is counted according to three-dimensional (3D stereologic counting rules by visual inspection within hundreds of microscopic fields-of-view per investigated brain or brain region. Reliance on visual inspection forces stereologic cell counting to be very labor-intensive and time-consuming, and is the main reason why biased, non-stereologic two-dimensional (2D cell counting approaches have remained in widespread use. We present an evaluation of the performance of modern automated cell detection and segmentation algorithms as a potential alternative to the manual approach in stereologic cell counting. The image data used in this study were 3D microscopic images of thick brain tissue sections prepared with a variety of commonly used nuclear and cytoplasmic stains. The evaluation compared the numbers and locations of cells identified unambiguously and counted exhaustively by an expert observer with those found by three automated 3D cell detection algorithms: nuclei segmentation from the FARSIGHT toolkit, nuclei segmentation by 3D multiple level set methods, and the 3D object counter plug-in for ImageJ. Of these methods, FARSIGHT performed best, with true-positive detection rates between 38–99% and false-positive rates from 3.6–82%. The results demonstrate that the current automated methods suffer from lower detection rates and higher false-positive rates than are acceptable for obtaining valid estimates of cell numbers. Thus, at present, stereologic cell counting with manual decision for object inclusion according to unbiased stereologic counting rules remains the only adequate method for unbiased cell quantification in histologic tissue sections.

  12. Short communication: Repeatability of differential goat bulk milk culture and associations with somatic cell count, total bacterial count, and standard plate count.

    Science.gov (United States)

    Koop, G; Dik, N; Nielen, M; Lipman, L J A

    2010-06-01

    The aims of this study were to assess how different bacterial groups in bulk milk are related to bulk milk somatic cell count (SCC), bulk milk total bacterial count (TBC), and bulk milk standard plate count (SPC) and to measure the repeatability of bulk milk culturing. On 53 Dutch dairy goat farms, 3 bulk milk samples were collected at intervals of 2 wk. The samples were cultured for SPC, coliform count, and staphylococcal count and for the presence of Staphylococcus aureus. Furthermore, SCC (Fossomatic 5000, Foss, Hillerød, Denmark) and TBC (BactoScan FC 150, Foss) were measured. Staphylococcal count was correlated to SCC (r=0.40), TBC (r=0.51), and SPC (r=0.53). Coliform count was correlated to TBC (r=0.33), but not to any of the other variables. Staphylococcus aureus did not correlate to SCC. The contribution of the staphylococcal count to the SPC was 31%, whereas the coliform count comprised only 1% of the SPC. The agreement of the repeated measurements was low. This study indicates that staphylococci in goat bulk milk are related to SCC and make a significant contribution to SPC. Because of the high variation in bacterial counts, repeated sampling is necessary to draw valid conclusions from bulk milk culturing. 2010 American Dairy Science Association. Published by Elsevier Inc. All rights reserved.

  13. Quantitative clinical uptake measurements using conjugate counting

    International Nuclear Information System (INIS)

    Lathrop, K.A.; Bartlett, R.D.; Chen, C.T.; Chou, J.S.; Faulhaber, P.F.; Harper, P.V.; Stark, V.J.

    1986-01-01

    While the use of conjugate counting for determination of organ uptake in human subjects has been extensively described, in the present study the determination of the organ uptake of ortho-iodohippurate presented several opportunities for validation of the in vivo counting data. Ortho-iodohippurate is distributed in the extracellular space, is largely extracted on each pass through the kidneys, and is not significantly deiodinated in vivo. Thus, the kidney uptake rate should be proportional to the blood level, the appearance rate of activity in the bladder is equal to the disappearance rate from the kidneys, and direct measurement of activity in the urine after voiding provides an internal standard for imaging measurements of bladder activity. Since the activity levels in the kidneys, bladder, and remainder of the body changed fairly rapidly, especially in the first 20 to 30 minutes following injection, posterior images of the trunk including kidneys and bladder were obtained continuously using a gamma camera fitted with a diverging collimator for 30 minutes and then at intervals for several hours. Simultaneous conjugate counting determinations were made using a whole body scanning system previously described at these meetings. Imaging data corrected for decay and adjacent background were fitted by least squares methods to curves representing a sum of exponentials, and the curves were normalized to the conjugate uptake measurements. The uptake curves of the kidneys and bladder matched well with the direct measurements of the urinary excretion. Data were collected in 16 normal subjects, and the estimated absorbed dose was calculated for the kidneys, the bladder and the remainder of the body for seven radioisotopes of iodine. 4 references, 6 figures, 2 tables

  14. Conversion of Beckman DK-2A spectrophotometer into an automatic single-photon counting fluorescence spectrophotometer

    International Nuclear Information System (INIS)

    Chikkur, G.C.; Lagare, M.T.; Umakantha, N.

    1981-01-01

    Details of how a DK-2A spectrophotometer can be modified into an automatic single-photon counting fluorescence spectrophotometer for recording a low intensity spectrum, are reported. The single-photon count-rate converted into a DC voltage is applied at the appropriate stage in the sample channel amplifier circuit of a DK-2A to get the pen deflection proportional to the count-rate. A high intensity spectrum may be recorded in the usual way by merely turning the shaft of the mirror motor by 180 degrees. (author)

  15. Development of a phoswich detector for neutron dose rate measurements in the Earth's atmosphere

    Energy Technology Data Exchange (ETDEWEB)

    Doensdorf, Esther Miriam

    2014-04-30

    of count rates and dose rates is determined. The count rates show a maximum which corresponds to the location of the Pfotzer maximum (Pfotzer, 1936). In the dose rates a maximum is found as well, which has been mentioned in Bartlett (2004).

  16. Method and apparstus for determining random coincidence count rate in a scintillation counter utilizing the coincidence technique

    International Nuclear Information System (INIS)

    Horrocks, D.L.

    1980-01-01

    A method and apparatus for the reliable determination of a random coincidence count attributable to chance coincidences of single-photon events which are each detected in only a single detector of a scintillation counter utilizing two detectors in a coincidence counting technique are described. A firstdelay device is employed to delay output pulses from one detector, and then the delayed signal is compared with the undelayed signal from the other detector in a coincidence circuit, to obtain an approximate random coincidence count. The output of the coincidence circuit is applied to an anti-coincidence circuit, where it is corrected by elimination of pulses coincident with, and attributable to, conventionally detected real coincidences, and by elimination of pulses coincident with, and attributable to, real coincidences that have been delayed by a second delay device having the same time parameter as the first. 8 claims

  17. A new method of quench monitoring in liquid scintillation counting

    International Nuclear Information System (INIS)

    Horrocks, D.L.

    1978-01-01

    The quench level of different liquid scintillation counting samples is measured by comparing the responses (pulse heights) produced by the same energy electrons in each sample. The electrons utilized in the measurements are those of the maximum energy (Esub(max)) which are produced by the single Compton scattering process for the same energy gamma-rays in each sample. The Esub(max) response produced in any sample is related to the Esub(max) response produced in an unquenched, sealed standard. The difference in response on a logarithm response scale is defined as the ''H Number''. The H number is related to the counting efficiency of the desired radionuclide by measurement of a set of standards of known amounts of the radionuclide and different amounts of quench (standard quench curve). The concept of the H number has been shown to be theoretically valid. Based upon this proof, the features of the H number concept as embodied in the Beckman LS-8000 Series Liquid Scintillation Systems have been demonstrated. It has been shown that one H number is unique; it provides a method of instrument calibration and wide dynamic quench range measurements. Further, it has been demonstrated that the H number concept provides a universal quench parameter. Counting efficiency vs. H number plots are repeatable within the statistical limits of +-1% counting efficiency. By the use of the H number concept a very accurate method of automatic quench compensation (A.Q.C.) is possible. (T.G.)

  18. Repeatability of differential goat bulk milk culture and associations with somatic cell count, total bacterial count, and standard plate count

    NARCIS (Netherlands)

    Koop, G.; Dik, N.; Nielen, M.; Lipman, L.J.A.

    2010-01-01

    The aims of this study were to assess how different bacterial groups in bulk milk are related to bulk milk somatic cell count (SCC), bulk milk total bacterial count (TBC), and bulk milk standard plate count (SPC) and to measure the repeatability of bulk milk culturing. On 53 Dutch dairy goat farms,

  19. Dose-rate determination by radiochemical analysis

    International Nuclear Information System (INIS)

    Mangini, A.; Pernicka, E.; Wagner, G.A.

    1983-01-01

    At the previous TL Specialist Seminr we had suggested that α-counting is an unsuitable technique for dose-rate determination due to overcounting effects. This is confirmed by combining α-counting, neutron activation analysis, fission track counting, α-spectrometry on various pottery samples. One result of this study is that disequilibrium in the uranium decay chain alone cannot account for the observed discrepancies between α-counting and chemical analysis. Therefore we propose for routine dose-rate determination in TL dating to apply chemical analysis of the radioactive elements supplemented by an α-spectrometric equilibrium check. (author)

  20. Correction for intrinsic and set dead-time losses in radioactivity counting

    International Nuclear Information System (INIS)

    Wyllie, H.A.

    1992-12-01

    Equations are derived for the determination of the intrinsic dead time of the components which precede the paralysis unit in a counting system for measuring radioactivity. The determination depends on the extension of the set dead time by the intrinsic dead time. Improved formulae are given for the dead-time correction of the count rate of a radioactive source in a single-channel system. A variable in the formulae is the intrinsic dead time which is determined concurrently with the counting of the source. The only extra equipment required in a conventional system is a scaler. 5 refs., 2 tabs., 21 figs

  1. Reduction in sperm count and increase in abnormal sperm in the mouse following x-irradiation or injection of 22Na

    International Nuclear Information System (INIS)

    Harrison, A.; Moore, P.C.

    1980-01-01

    The accumulated reduction in sperm count and increase in the number of aberrant sperm in the mouse were used to compare acute X-irradiation with protracted radiation from an injection of 22 Na. The effects per unit of absorbed dose of acute and protracted radiation were similar on sperm count, but in respect of numbers of abnormal sperm the 22 Na induced a significantly greater maximum than X-rays. (author)

  2. Comparative analysis of dose rates in bricks determined by neutron activation analysis, alpha counting and X-ray fluorescence analysis for the thermoluminescence fine grain dating method

    Science.gov (United States)

    Bártová, H.; Kučera, J.; Musílek, L.; Trojek, T.

    2014-11-01

    In order to evaluate the age from the equivalent dose and to obtain an optimized and efficient procedure for thermoluminescence (TL) dating, it is necessary to obtain the values of both the internal and the external dose rates from dated samples and from their environment. The measurements described and compared in this paper refer to bricks from historic buildings and a fine-grain dating method. The external doses are therefore negligible, if the samples are taken from a sufficient depth in the wall. However, both the alpha dose rate and the beta and gamma dose rates must be taken into account in the internal dose. The internal dose rate to fine-grain samples is caused by the concentrations of natural radionuclides 238U, 235U, 232Th and members of their decay chains, and by 40K concentrations. Various methods can be used for determining trace concentrations of these natural radionuclides and their contributions to the dose rate. The dose rate fraction from 238U and 232Th can be calculated, e.g., from the alpha count rate, or from the concentrations of 238U and 232Th, measured by neutron activation analysis (NAA). The dose rate fraction from 40K can be calculated from the concentration of potassium measured, e.g., by X-ray fluorescence analysis (XRF) or by NAA. Alpha counting and XRF are relatively simple and are accessible for an ordinary laboratory. NAA can be considered as a more accurate method, but it is more demanding regarding time and costs, since it needs a nuclear reactor as a neutron source. A comparison of these methods allows us to decide whether the time- and cost-saving simpler techniques introduce uncertainty that is still acceptable.

  3. Standardization of {sup 241}Am by digital coincidence counting, liquid scintillation counting and defined solid angle counting

    Energy Technology Data Exchange (ETDEWEB)

    Balpardo, C., E-mail: balpardo@cae.cnea.gov.a [Laboratorio de Metrologia de Radioisotopos, CNEA, Buenos Aires (Argentina); Capoulat, M.E.; Rodrigues, D.; Arenillas, P. [Laboratorio de Metrologia de Radioisotopos, CNEA, Buenos Aires (Argentina)

    2010-07-15

    The nuclide {sup 241}Am decays by alpha emission to {sup 237}Np. Most of the decays (84.6%) populate the excited level of {sup 237}Np with energy of 59.54 keV. Digital coincidence counting was applied to standardize a solution of {sup 241}Am by alpha-gamma coincidence counting with efficiency extrapolation. Electronic discrimination was implemented with a pressurized proportional counter and the results were compared with two other independent techniques: Liquid scintillation counting using the logical sum of double coincidences in a TDCR array and defined solid angle counting taking into account activity inhomogeneity in the active deposit. The results show consistency between the three methods within a limit of a 0.3%. An ampoule of this solution will be sent to the International Reference System (SIR) during 2009. Uncertainties were analysed and compared in detail for the three applied methods.

  4. CERN{sub D}xCTA counting mode chip

    Energy Technology Data Exchange (ETDEWEB)

    Moraes, D. [CERN, CH-1211 Geneva 23 (Switzerland)], E-mail: danielle.moraes@cern.ch; Kaplon, J. [CERN, CH-1211 Geneva 23 (Switzerland); Nygard, E. [Interon AS, Asker, Norway and DX-ray Inc., Northridge, CA (United States)

    2008-06-11

    This ASIC is a counting mode front-end electronic optimized for the readout of CdZnTe/CdTe and silicon sensors, for possible use in applications where the flux of ionizing radiation is high. The chip is implemented in 0.25 {mu}m CMOS technology. The circuit comprises 128 channels equipped with a transimpedance amplifier followed by a gain shaper stage with 21 ns peaking time, two discriminators and two 18-bit counters. The channel architecture is optimized for the detector characteristics in order to achieve the best energy resolution at counting rates of up to 5 M counts/second. The amplifier shows a linear sensitivity of 118 mV/fC and an equivalent noise charge of about 711 e{sup -}, for a detector capacitance of 5 pF. Complete evaluation of the circuit is presented using electronic pulses and pixel detectors.

  5. A fast, dataflow-oriented preprocessor for cluster counting

    International Nuclear Information System (INIS)

    Imhof, M.

    1995-01-01

    One of the main problems in the use of cluster counting for particle identification in high energy physics is the huge primary data rate. By means of an online data reduction system, this data rate can be reduced to the amount known from conventional integral charge/time readout. For tests and optimization of online data reduction algorithms, a RISC-based, data-flow-oriented multiprocessor board has been designed. (orig.)

  6. Gas detectors for thermal neutron at high counting rates; Etude des detecteurs a gaz pour neutrons thermiques fonctionnant en collection de courant

    Energy Technology Data Exchange (ETDEWEB)

    Mai, V Q [Commissariat a l' Energie Atomique, Grenoble (France). Centre d' Etudes Nucleaires

    1965-09-01

    After reminding of the current pulse formation theory in a cylindrical shape counter with and without gas multiplication, one gives the schemes of pulse amplifier and level discriminator which have allowed to verify the above calculations and to make clear the part at high counting rates of the space charge in proportional counters. The theory of that phenomenon is given in chapter V I at last, one gives the results obtained in a nuclear reactor with a counting-channel built with the above electronics circuits. (author) [French] Apres avoir rappele la theorie de la formation de l'impulsion de courant dans un compteur a geometrie cylindrique, fonctionnant sans et avec multiplication de charge, on etudie l'amplificateur et le discriminateur qui ont permis de verifier experimentalement les calculs precedents et de mettre en evidence l'action de la charge d'espace dans les compteurs proportionnels fonctionnant a tres fort taux de comptage. Une theorie de ce phenomene est donnee au chapitre VI; on indique enfin les resultats obtenus dans un reacteur avec une chaine de comptage utilisant les circuits electroniques precedents. (auteur)

  7. Daily step count predicts acute exacerbations in a US cohort with COPD.

    Directory of Open Access Journals (Sweden)

    Marilyn L Moy

    Full Text Available BACKGROUND: COPD is characterized by variability in exercise capacity and physical activity (PA, and acute exacerbations (AEs. Little is known about the relationship between daily step count, a direct measure of PA, and the risk of AEs, including hospitalizations. METHODS: In an observational cohort study of 169 persons with COPD, we directly assessed PA with the StepWatch Activity Monitor, an ankle-worn accelerometer that measures daily step count. We also assessed exercise capacity with the 6-minute walk test (6MWT and patient-reported PA with the St. George's Respiratory Questionnaire Activity Score (SGRQ-AS. AEs and COPD-related hospitalizations were assessed and validated prospectively over a median of 16 months. RESULTS: Mean daily step count was 5804±3141 steps. Over 209 person-years of observation, there were 263 AEs (incidence rate 1.3±1.6 per person-year and 116 COPD-related hospitalizations (incidence rate 0.56±1.09 per person-year. Adjusting for FEV1 % predicted and prednisone use for AE in previous year, for each 1000 fewer steps per day walked at baseline, there was an increased rate of AEs (rate ratio 1.07; 95%CI = 1.003-1.15 and COPD-related hospitalizations (rate ratio 1.24; 95%CI = 1.08-1.42. There was a significant linear trend of decreasing daily step count by quartiles and increasing rate ratios for AEs (P = 0.008 and COPD-related hospitalizations (P = 0.003. Each 30-meter decrease in 6MWT distance was associated with an increased rate ratio of 1.07 (95%CI = 1.01-1.14 for AEs and 1.18 (95%CI = 1.07-1.30 for COPD-related hospitalizations. Worsening of SGRQ-AS by 4 points was associated with an increased rate ratio of 1.05 (95%CI = 1.01-1.09 for AEs and 1.10 (95%CI = 1.02-1.17 for COPD-related hospitalizations. CONCLUSIONS: Lower daily step count, lower 6MWT distance, and worse SGRQ-AS predict future AEs and COPD-related hospitalizations, independent of pulmonary function and previous AE

  8. Detection and Counting of Orchard Trees from Vhr Images Using a Geometrical-Optical Model and Marked Template Matching

    Science.gov (United States)

    Maillard, Philippe; Gomes, Marília F.

    2016-06-01

    This article presents an original algorithm created to detect and count trees in orchards using very high resolution images. The algorithm is based on an adaptation of the "template matching" image processing approach, in which the template is based on a "geometricaloptical" model created from a series of parameters, such as illumination angles, maximum and ambient radiance, and tree size specifications. The algorithm is tested on four images from different regions of the world and different crop types. These images all have the GoogleEarth application. Results show that the algorithm is very efficient at detecting and counting trees as long as their spectral and spatial characteristics are relatively constant. For walnut, mango and orange trees, the overall accuracy was clearly above 90%. However, the overall success rate for apple trees fell under 75%. It appears that the openness of the apple tree crown is most probably responsible for this poorer result. The algorithm is fully explained with a step-by-step description. At this stage, the algorithm still requires quite a bit of user interaction. The automatic determination of most of the required parameters is under development.

  9. Logical error rate scaling of the toric code

    International Nuclear Information System (INIS)

    Watson, Fern H E; Barrett, Sean D

    2014-01-01

    To date, a great deal of attention has focused on characterizing the performance of quantum error correcting codes via their thresholds, the maximum correctable physical error rate for a given noise model and decoding strategy. Practical quantum computers will necessarily operate below these thresholds meaning that other performance indicators become important. In this work we consider the scaling of the logical error rate of the toric code and demonstrate how, in turn, this may be used to calculate a key performance indicator. We use a perfect matching decoding algorithm to find the scaling of the logical error rate and find two distinct operating regimes. The first regime admits a universal scaling analysis due to a mapping to a statistical physics model. The second regime characterizes the behaviour in the limit of small physical error rate and can be understood by counting the error configurations leading to the failure of the decoder. We present a conjecture for the ranges of validity of these two regimes and use them to quantify the overhead—the total number of physical qubits required to perform error correction. (paper)

  10. Diagnostic value of inflammatory markers (complete blood count ...

    African Journals Online (AJOL)

    Objective The aim of the study was to evaluate the diagnostic value of inflammatory markers [complete blood cell count (CBC), erythrocyte sedimentation rate (ESR),. C-reactive protein (CRP)] for the differentiation of acute appendicitis from nonspecific abdominal pain in children. Patients and methods In this prospective ...

  11. Diagnostic value of inflammatory markers (complete blood count ...

    African Journals Online (AJOL)

    Objective The aim of the study was to evaluate the diagnostic value of inflammatory markers [complete blood cell count (CBC), erythrocyte sedimentation rate (ESR), C-reactive protein (CRP)] for the differentiation of acute appendicitis from nonspecific abdominal pain in children. Patients and methods In this prospective ...

  12. Quantitative monitoring of the fluorination process by neutron counting

    International Nuclear Information System (INIS)

    Russo, P.A.; Appert, Q.D.; Biddle, R.S.; Kelley, T.A.; Martinez, M.M.; West, M.H.

    1993-01-01

    Plutonium metal is produced by reducing PuF 4 prepared from PuO 2 by fluorination. Both fluorination and reduction are batch processes at the Los Alamos Plutonium Facility. The conversion of plutonium oxide to fluoride greatly increases the neutron yield, a result of the high cross section for alpha-neutron (α,n) reactions on fluorine targets compared to the (more than 100 times) smaller α,n yield on oxygen targets. Because of the increase, total neutron counting can be used to monitor the conversion process. This monitoring ability can lead to an improved metal product, reduced scrap for recycle, waste reduction, minimized reagent usage, and reduce personnel radiation exposures. A new stirred-bed fluorination process has been developed simultaneously with a recent evaluation of an automated neutron-counting instrument for quantitative process monitoring. Neutrons are counted with polyethylene-moderated 3 He-gas proportional counters. Results include a calibration of the real-time neutron-count-rate indicator for the extent of fluorination using reference values obtained from destructive analysis of samples from the blended fluoroinated batch

  13. Clean Hands Count

    Medline Plus

    Full Text Available ... Like this video? Sign in to make your opinion count. Sign in 131 2 Don't like this video? Sign in to make your opinion count. Sign in 3 Loading... Loading... Transcript The ...

  14. An Economical Fast Discriminator for Nuclear Pulse Counting

    International Nuclear Information System (INIS)

    Issarachai, Opas; Punnachaiya, Suvit

    2009-07-01

    Full text: This research work was aimed to develop a fast discriminator at low cost but high capability for discrimination a nanosecond nuclear pulse. The fast discriminator can be used in association with fast photon counting system. The designed structure consisted of the ultra-fast voltage comparator using ADCMP601 integrated circuit, the monostable multivibrator with controllable pulse width output by propagation delay of logic gate, and the fast response buffer amplifier. The tested results of pulse height discrimination of 0-5 V nuclear pulse with 20 ns (FWHM) pulse width showed the correlation coefficient (R 2 ) between discrimination level and pulse height was 0.998, while the pulse rate more than 10 MHz could be counted. The 30 ns logic pulse width output revealed high stable and could be smoothly driven to low impedance load at 50 Ω. For pulse signal transmission to the counter, it was also found that the termination of reflected signal must be considered because it may cause pulse counting error

  15. Characterization of imaging pixel detectors of Si and CdTe read out with the counting X-ray chip MPEC 2.3

    International Nuclear Information System (INIS)

    Loecker, M.

    2007-04-01

    Single photon counting detectors with Si- and CdTe-sensors have been constructed and characterized. As readout chip the MPEC 2.3 is used which consists of 32 x 32 pixels with 200 x 200 μm 2 pixel size and which has a high count rate cabability (1 MHz per pixel) as well as a low noise performance (55 e - ). Measurements and simulations of the detector homogeneity are presented. It could be shown that the theoretical maximum of the homogeneity is reached (quantum limit). By means of the double threshold of the MPEC chip the image contrast can be enhanced which is demonstrated by measurement and simulation. Also, multi-chip-modules consisting of 4 MPEC chips and a single Si- or CdTe-sensor have been constructed and successfully operated. With these modules modulation-transfer-function measurements have been done showing a good spatial resolution of the detectors. In addition, multi-chip-modules according to the Sparse-CMOS concept have been built and tests characterizing the interconnection technologies have been performed

  16. Musculoskeletal imaging with a prototype photon-counting detector.

    Science.gov (United States)

    Gruber, M; Homolka, P; Chmeissani, M; Uffmann, M; Pretterklieber, M; Kainberger, F

    2012-01-01

    To test a digital imaging X-ray device based on the direct capture of X-ray photons with pixel detectors, which are coupled with photon-counting readout electronics. The chip consists of a matrix of 256 × 256 pixels with a pixel pitch of 55 μm. A monolithic image of 11.2 cm × 7 cm was obtained by the consecutive displacement approach. Images of embalmed anatomical specimens of eight human hands were obtained at four different dose levels (skin dose 2.4, 6, 12, 25 μGy) with the new detector, as well as with a flat-panel detector. The overall rating scores for the evaluated anatomical regions ranged from 5.23 at the lowest dose level, 6.32 at approximately 6 μGy, 6.70 at 12 μGy, to 6.99 at the highest dose level with the photon-counting system. The corresponding rating scores for the flat-panel detector were 3.84, 5.39, 6.64, and 7.34. When images obtained at the same dose were compared, the new system outperformed the conventional DR system at the two lowest dose levels. At the higher dose levels, there were no significant differences between the two systems. The photon-counting detector has great potential to obtain musculoskeletal images of excellent quality at very low dose levels.

  17. Asthma and myocardial infarction inpatient hospitalization and emergency room visit counts and rates by county, year and month of admission, age group, race/ethnicity and gender of California residents, 2000-2009.

    Data.gov (United States)

    California Environmental Health Tracking Program — This dataset contains case counts, rates, and confidence intervals of asthma (ICD9-CM 493.0-493.9) and myocardial infarction (ICD9-CM 410) inpatient hospitalizations...

  18. Noun Countability; Count Nouns and Non-count Nouns, What are the Syntactic Differences Between them?

    Directory of Open Access Journals (Sweden)

    Azhar A. Alkazwini

    2016-11-01

    Full Text Available Words that function as the subjects of verbs, objects of verbs or prepositions and which can have a plural form and possessive ending are known as nouns. They are described as referring to persons, places, things, states, or qualities and might also be used as an attributive modifier. In this paper, classes and subclasses of nouns shall be presented, then, noun countability branching into count and non-count nous shall be discussed. A number of present examples illustrating differences between count and non-count nouns and this includes determiner-head-co-occurrence restrictions of number, subject-verb agreement, in addition to some exceptions to this agreement rule shall be discussed. Also, the lexically inherent number in nouns and how inherently plural nouns are classified in terms of (+/- count are illustrated. This research will discuss partitive construction of count and non-count nouns, nouns as attributive modifier and, finally, conclude with the fact that there are syntactic difference between count and non-count in the English Language.

  19. Eosinophil count is positively correlated with coronary artery calcification

    International Nuclear Information System (INIS)

    Tanaka, Muhei; Fukui, Michiaki; Yamasaki, Masahiro; Hasegawa, Goji; Oda, Yohei; Nakamura, Naoto; Tomiyasu, Ki-ichiro; Akabame, Satoshi; Nakano, Koji

    2012-01-01

    Recent studies suggested that allergic disorders and increased eosinophil count were associated with atherosclerosis. The purpose of this study was to assess the relationship between eosinophil count and coronary artery calcification (CAC). We performed a cross-sectional study in 1363 consecutive participants with clinical suspicion of coronary heart disease (CHD). We evaluated the relationships between CAC score determined by multislice CT and peripheral eosinophil count as well as major cardiovascular risk factors, including age, body mass index, smoking status, hypertension, dyslipidemia, diabetes mellitus (DM), high-sensitivity C-reactive protein and estimated glomerular filtration rate (eGFR). Sex (P=0.0004), hypertension (P=0.0002), dyslipidemia (P=0.0004) and DM (P=0.0061) were associated with log (CAC+1), respectively. Positive correlations were found between log (CAC+1), and age (r=0.325, P<0.0001) and eosinophil count (r=0.165, P<0.0001). Negative correlations were found between log (CAC+1) and eGFR (r=-0.166, P<0.0001). Multivariate linear regression analysis demonstrated that age (β=0.314, P<0.0001), sex (β=0.124, P<0.0001), hypertension (β=0.084, P=0.0008), DM (β=0.108, P<0.0001), eGFR (β=-0.079, P=0.0021) and eosinophil count (β=0.147, P<0.0001) were independent determinants of log (CAC+1). In conclusion, eosinophil count correlated positively with CAC in participants with clinical suspicion of CHD. (author)

  20. Single ion counting with a MCP (microchannel plate) detector

    Energy Technology Data Exchange (ETDEWEB)

    Tawara, Hiroko; Sasaki, Shinichi; Miyajima, Mitsuhiro [National Lab. for High Energy Physics, Tsukuba, Ibaraki (Japan); Shibamura, Eido

    1996-07-01

    In this study, a single-ion-counting method using alpha-particle-impact ionization of Ar atoms is demonstrated and the preliminary {epsilon}{sub mcp} for Ar ions with incident energies of 3 to 4.7 keV is determined. The single-ion counting by the MCP is aimed to be performed under experimental conditions as follows: (1) A signal from the MCP is reasonably identified as incidence of single Ar-ion. (2) The counting rate of Ar ions is less than 1 s{sup -1}. (3) The incident Ar ions are not focused on a small part of an active area of the MCP, namely, {epsilon}{sub mcp} is determined with respect to the whole active area of the MCP. So far, any absolute detection efficiency has not been reported under these conditions. (J.P.N.)

  1. Clean Hands Count

    Medline Plus

    Full Text Available ... starting stop Loading... Watch Queue Queue __count__/__total__ It’s YouTube. Uninterrupted. Loading... Want music and videos with ... ads? Get YouTube Red. Working... Not now Try it free Find out why Close Clean Hands Count ...

  2. Maximum type 1 error rate inflation in multiarmed clinical trials with adaptive interim sample size modifications.

    Science.gov (United States)

    Graf, Alexandra C; Bauer, Peter; Glimm, Ekkehard; Koenig, Franz

    2014-07-01

    Sample size modifications in the interim analyses of an adaptive design can inflate the type 1 error rate, if test statistics and critical boundaries are used in the final analysis as if no modification had been made. While this is already true for designs with an overall change of the sample size in a balanced treatment-control comparison, the inflation can be much larger if in addition a modification of allocation ratios is allowed as well. In this paper, we investigate adaptive designs with several treatment arms compared to a single common control group. Regarding modifications, we consider treatment arm selection as well as modifications of overall sample size and allocation ratios. The inflation is quantified for two approaches: a naive procedure that ignores not only all modifications, but also the multiplicity issue arising from the many-to-one comparison, and a Dunnett procedure that ignores modifications, but adjusts for the initially started multiple treatments. The maximum inflation of the type 1 error rate for such types of design can be calculated by searching for the "worst case" scenarios, that are sample size adaptation rules in the interim analysis that lead to the largest conditional type 1 error rate in any point of the sample space. To show the most extreme inflation, we initially assume unconstrained second stage sample size modifications leading to a large inflation of the type 1 error rate. Furthermore, we investigate the inflation when putting constraints on the second stage sample sizes. It turns out that, for example fixing the sample size of the control group, leads to designs controlling the type 1 error rate. © 2014 The Author. Biometrical Journal published by WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  3. Releasable activity and maximum permissible leakage rate within a transport cask of Tehran Research Reactor fuel samples

    Directory of Open Access Journals (Sweden)

    Rezaeian Mahdi

    2015-01-01

    Full Text Available Containment of a transport cask during both normal and accident conditions is important to the health and safety of the public and of the operators. Based on IAEA regulations, releasable activity and maximum permissible volumetric leakage rate within the cask containing fuel samples of Tehran Research Reactor enclosed in an irradiated capsule are calculated. The contributions to the total activity from the four sources of gas, volatile, fines, and corrosion products are treated separately. These calculations are necessary to identify an appropriate leak test that must be performed on the cask and the results can be utilized as the source term for dose evaluation in the safety assessment of the cask.

  4. Count-to-count time interval distribution analysis in a fast reactor

    International Nuclear Information System (INIS)

    Perez-Navarro Gomez, A.

    1973-01-01

    The most important kinetic parameters have been measured at the zero power fast reactor CORAL-I by means of the reactor noise analysis in the time domain, using measurements of the count-to-count time intervals. (Author) 69 refs

  5. Maximum Evaporation Rates of Water Droplets Approaching Obstacles in the Atmosphere Under Icing Conditions

    Science.gov (United States)

    Lowell, H. H.

    1953-01-01

    When a closed body or a duct envelope moves through the atmosphere, air pressure and temperature rises occur ahead of the body or, under ram conditions, within the duct. If cloud water droplets are encountered, droplet evaporation will result because of the air-temperature rise and the relative velocity between the droplet and stagnating air. It is shown that the solution of the steady-state psychrometric equation provides evaporation rates which are the maximum possible when droplets are entrained in air moving along stagnation lines under such conditions. Calculations are made for a wide variety of water droplet diameters, ambient conditions, and flight Mach numbers. Droplet diameter, body size, and Mach number effects are found to predominate, whereas wide variation in ambient conditions are of relatively small significance in the determination of evaporation rates. The results are essentially exact for the case of movement of droplets having diameters smaller than about 30 microns along relatively long ducts (length at least several feet) or toward large obstacles (wings), since disequilibrium effects are then of little significance. Mass losses in the case of movement within ducts will often be significant fractions (one-fifth to one-half) of original droplet masses, while very small droplets within ducts will often disappear even though the entraining air is not fully stagnated. Wing-approach evaporation losses will usually be of the order of several percent of original droplet masses. Two numerical examples are given of the determination of local evaporation rates and total mass losses in cases involving cloud droplets approaching circular cylinders along stagnation lines. The cylinders chosen were of 3.95-inch (10.0+ cm) diameter and 39.5-inch 100+ cm) diameter. The smaller is representative of icing-rate measurement cylinders, while with the larger will be associated an air-flow field similar to that ahead of an airfoil having a leading-edge radius

  6. Absorbed dose rate meter for β-ray

    International Nuclear Information System (INIS)

    Bingo, K.

    1977-01-01

    The absorbed dose of β-ray depends on the energy of β-rays and the epidermal thickness of tissue in interest. In order to measure the absorbed dose rate at the interested tissue directly, the ratio of counting rate to absorbed dose should be constant independent of β-ray energy. In this purpose, a thin plastic scintillator was used as a detector with a single channel analyzer. The pulse height distribution, obtained using the scintillator whose thickness is less than the range of β-rays, shows a peak at a particular pulse height depending on the thickness of scintillator used. This means an increase of the number of pulses at lower pulse height. The lower level of discrimination and window width of the single channel analyzer are chosen according to the epidermal thickness of the tissue. In the experiment, scintillators of 0.5, 1, 2, 3, 5 and 10 mm thick were tested. It was found that desirable pulse height distribution, to obtain a constant dose sensitivity, could be obtained using the 2 mm thick scintillator. The sensitivity of the absorbed dose rate meter is constant within +-15% for β-ray with maximum energy from 0.4 to 3.5 MeV, when the absorbed dose rate for skin (epidermal thickness 7mg/cm 2 ) is measured. In order to measure the dose rate for a hand (epithermal thickness 40mg/cm 2 ) the lower level of discrimination is changed to be higher and at the same time the window width is also changed. Combining these techniques, one can get an absorbed dose rate meter for the tissue dose of various thickness, which has the constant dose sensitivity within +-15% for β-rays with maximum energy from 0.4 to 3.5 MeV

  7. 78 FR 13999 - Maximum Interest Rates on Guaranteed Farm Loans

    Science.gov (United States)

    2013-03-04

    ..., cost-plus, flat-rate, or market based) to price guaranteed loans, provided the rates do not exceed the... (LIBOR) or the 5-year Treasury note rate, unless the lender uses a formal written risk-based pricing... cost in the form of a lower interest rate than the borrower would otherwise receive. Therefore, the FSA...

  8. HIV Seropositivity And CD4 T-Lymphocyte Counts Among Infants ...

    African Journals Online (AJOL)

    The CD4 cell count was estimated using the Dynamal ® Quant Kit (Dynal Biotechn, ASA, Oslo, Norway). Results: The overall HIV prevalence rate in this study was 23.2%. The distribution of HIV prevalence among different age group revealed a high prevalence rate among the under fives (24.1% for males and 26.4% for ...

  9. DAWN GRAND MAP CERES TPE NEUTRON COUNTS V1.0

    Data.gov (United States)

    National Aeronautics and Space Administration — A global map thermal+epithermal neutron counting rates binned on twenty-degree quasi-equal-area pixels is provided. The map was determined from a time series of the...

  10. Estimation of the players maximum heart rate in real game situations in team sports: a practical propose

    Directory of Open Access Journals (Sweden)

    Jorge Cuadrado Reyes

    2011-05-01

    Full Text Available Abstract   This  research developed a logarithms  for calculating the maximum heart rate (max. HR for players in team sports in  game situations. The sample was made of  thirteen players (aged 24 ± 3   to a  Division Two Handball team. HR was initially measured by Course Navette test.  Later, twenty one training sessions were conducted  in which HR and Rate of Perceived Exertion (RPE, were  continuously monitored, in each task. A lineal regression analysis was done  to help find a max. HR prediction equation from the max. HR of the three highest intensity sessions. Results from  this equation correlate significantly with data obtained in the Course Navette test and with those obtained by other indirect methods. The conclusion of this research is that this equation provides a very useful and easy way to measure the max. HR in real game situations, avoiding non-specific analytical tests and, therefore laboratory testing..   Key words: workout control, functional evaluation, prediction equation.

  11. Correction of count losses due to deadtime on a DST-XLi (SMVi-GE) camera during dosimetric studies in patients injected with iodine-131

    International Nuclear Information System (INIS)

    Delpon, G.; Ferrer, L.; Lisbona, A.; Bardies, M.

    2002-01-01

    In dosimetric studies performed after therapeutic injection, it is essential to correct count losses due to deadtime on the gamma camera. This note describes four deadtime correction methods, one based on the use of a standard source without preliminary calibration, and three requiring specific calibration and based on the count rate observed in different spectrometric windows (20%, 20% plus a lower energy window and the full spectrum of 50-750 keV). Experiments were conducted on a phantom at increasingly higher count rates to check correction accuracy with the different methods. The error was less than +7% with a standard source, whereas count-rate-based methods gave more accurate results. On the assumption that the model was paralysable, preliminary calibration allowed an observed count rate curve to be plotted as a function of the real count rate. The use of the full spectrum led to a 3.0% underestimation for the highest activity imaged. As count losses depend on photon flux independent of energy, the use of the full spectrum during measurement allowed scatter conditions to be taken into account. A protocol was developed to apply this correction method to whole-body acquisitions. (author)

  12. An examination of the time-dependent background counts of the delayed neutron counting system at the Royal Military College of Canada

    International Nuclear Information System (INIS)

    Sellers, M.T.; Corcoran, E.C.; Kelly, D.G.

    2011-01-01

    A delayed neutron counting (DNC) system for the analysis of special nuclear materials (SNM) has been constructed and calibrated at the Royal Military College of Canada. The polyethylene vials used to transport SNM samples have been found to contribute a time-dependent count rate, B(t), far above the system background. B(t) has been found to be independent of polyethylene mass and shows a dependence on irradiation position in the SLOWPOKE-2 reactor and irradiation time. A comparison of B(t) and the theoretical delayed neutron production from the fission of small amounts of 235 U has indicated that trace amounts of uranium may be present in the DNC system tubing. (author)

  13. How Fred Hoyle Reconciled Radio Source Counts and the Steady State Cosmology

    Science.gov (United States)

    Ekers, Ron

    2012-09-01

    In 1969 Fred Hoyle invited me to his Institute of Theoretical Astronomy (IOTA) in Cambridge to work with him on the interpretation of the radio source counts. This was a period of extreme tension with Ryle just across the road using the steep slope of the radio source counts to argue that the radio source population was evolving and Hoyle maintaining that the counts were consistent with the steady state cosmology. Both of these great men had made some correct deductions but they had also both made mistakes. The universe was evolving, but the source counts alone could tell us very little about cosmology. I will try to give some indication of the atmosphere and the issues at the time and look at what we can learn from this saga. I will conclude by briefly summarising the exponential growth of the size of the radio source counts since the early days and ask whether our understanding has grown at the same rate.

  14. Characterization of spectrometric photon-counting X-ray detectors at different pitches

    Science.gov (United States)

    Jurdit, M.; Brambilla, A.; Moulin, V.; Ouvrier-Buffet, P.; Radisson, P.; Verger, L.

    2017-09-01

    There is growing interest in energy-sensitive photon-counting detectors based on high flux X-ray imaging. Their potential applications include medical imaging, non-destructive testing and security. Innovative detectors of this type will need to count individual photons and sort them into selected energy bins, at several million counts per second and per mm2. Cd(Zn)Te detector grade materials with a thickness of 1.5 to 3 mm and pitches from 800 μm down to 200 μm were assembled onto interposer boards. These devices were tested using in-house-developed full-digital fast readout electronics. The 16-channel demonstrators, with 256 energy bins, were experimentally characterized by determining spectral resolution, count rate, and charge sharing, which becomes challenging at low pitch. Charge sharing correction was found to efficiently correct X-ray spectra up to 40 × 106 incident photons.s-1.mm-2.

  15. The effect of volume and quenching on estimation of counting efficiencies in liquid scintillation counting

    International Nuclear Information System (INIS)

    Knoche, H.W.; Parkhurst, A.M.; Tam, S.W.

    1979-01-01

    The effect of volume on the liquid scintillation counting performance of 14 C-samples has been investigated. A decrease in counting efficiency was observed for samples with volumes below about 6 ml and those above about 18 ml when unquenched samples were assayed. Two quench-correction methods, sample channels ratio and external standard channels ratio, and three different liquid scintillation counters, were used in an investigation to determine the magnitude of the error in predicting counting efficiencies when small volume samples (2 ml) with different levels of quenching were assayed. The 2 ml samples exhibited slightly greater standard deviations of the difference between predicted and determined counting efficiencies than did 15 ml samples. Nevertheless, the magnitude of the errors indicate that if the sample channels ratio method of quench correction is employed, 2 ml samples may be counted in conventional counting vials with little loss in counting precision. (author)

  16. Estimation of DMFT, Salivary Streptococcus Mutans Count, Flow Rate, Ph, and Salivary Total Calcium Content in Pregnant and Non-Pregnant Women: A Prospective Study.

    Science.gov (United States)

    Kamate, Wasim Ismail; Vibhute, Nupura Aniket; Baad, Rajendra Krishna

    2017-04-01

    Pregnancy, a period from conception till birth, causes changes in the functioning of the human body as a whole and specifically in the oral cavity that may favour the emergence of dental caries. Many studies have shown pregnant women at increased risk for dental caries, however, specific salivary caries risk factors and the particular period of pregnancy at heightened risk for dental caries are yet to be explored and give a scope of further research in this area. The aim of the present study was to assess the severity of dental caries in pregnant women compared to non-pregnant women by evaluating parameters like Decayed, Missing, Filled Teeth (DMFT) index, salivary Streptococcus mutans count, flow rate, pH and total calcium content. A total of 50 first time pregnant women in the first trimester were followed during their second trimester, third trimester and postpartum period for the evaluation of DMFT by World Health Organization (WHO) scoring criteria, salivary flow rate by drooling method, salivary pH by pH meter, salivary total calcium content by bioassay test kit and salivary Streptococcus mutans count by semiautomatic counting of colonies grown on Mitis Salivarius (MS) agar supplemented by 0.2U/ml of bacitracin and 10% sucrose. The observations of pregnant women were then compared with same parameters evaluated in the 50 non-pregnant women. Paired t-test and Wilcoxon sign rank test were performed to assess the association between the study parameters. Evaluation of different caries risk factors between pregnant and non-pregnant women clearly showed that pregnant women were at a higher risk for dental caries. Comparison of caries risk parameters during the three trimesters and postpartum period showed that the salivary Streptococcus mutans count had significantly increased in the second trimester , third trimester and postpartum period while the mean pH and mean salivary total calcium content decreased in the third trimester and postpartum period. These

  17. Estimation of DMFT, Salivary Streptococcus Mutans Count, Flow Rate, Ph, and Salivary Total Calcium Content in Pregnant and Non-Pregnant Women: A Prospective Study

    Science.gov (United States)

    Vibhute, Nupura Aniket; Baad, Rajendra Krishna

    2017-01-01

    Introduction Pregnancy, a period from conception till birth, causes changes in the functioning of the human body as a whole and specifically in the oral cavity that may favour the emergence of dental caries. Many studies have shown pregnant women at increased risk for dental caries, however, specific salivary caries risk factors and the particular period of pregnancy at heightened risk for dental caries are yet to be explored and give a scope of further research in this area. Aim The aim of the present study was to assess the severity of dental caries in pregnant women compared to non-pregnant women by evaluating parameters like Decayed, Missing, Filled Teeth (DMFT) index, salivary Streptococcus mutans count, flow rate, pH and total calcium content. Materials and Methods A total of 50 first time pregnant women in the first trimester were followed during their second trimester, third trimester and postpartum period for the evaluation of DMFT by World Health Organization (WHO) scoring criteria, salivary flow rate by drooling method, salivary pH by pH meter, salivary total calcium content by bioassay test kit and salivary Streptococcus mutans count by semiautomatic counting of colonies grown on Mitis Salivarius (MS) agar supplemented by 0.2U/ml of bacitracin and 10% sucrose. The observations of pregnant women were then compared with same parameters evaluated in the 50 non-pregnant women. Paired t-test and Wilcoxon sign rank test were performed to assess the association between the study parameters. Results Evaluation of different caries risk factors between pregnant and non-pregnant women clearly showed that pregnant women were at a higher risk for dental caries. Comparison of caries risk parameters during the three trimesters and postpartum period showed that the salivary Streptococcus mutans count had significantly increased in the second trimester, third trimester and postpartum period while the mean pH and mean salivary total calcium content decreased in the third

  18. TasselNet: counting maize tassels in the wild via local counts regression network.

    Science.gov (United States)

    Lu, Hao; Cao, Zhiguo; Xiao, Yang; Zhuang, Bohan; Shen, Chunhua

    2017-01-01

    Accurately counting maize tassels is important for monitoring the growth status of maize plants. This tedious task, however, is still mainly done by manual efforts. In the context of modern plant phenotyping, automating this task is required to meet the need of large-scale analysis of genotype and phenotype. In recent years, computer vision technologies have experienced a significant breakthrough due to the emergence of large-scale datasets and increased computational resources. Naturally image-based approaches have also received much attention in plant-related studies. Yet a fact is that most image-based systems for plant phenotyping are deployed under controlled laboratory environment. When transferring the application scenario to unconstrained in-field conditions, intrinsic and extrinsic variations in the wild pose great challenges for accurate counting of maize tassels, which goes beyond the ability of conventional image processing techniques. This calls for further robust computer vision approaches to address in-field variations. This paper studies the in-field counting problem of maize tassels. To our knowledge, this is the first time that a plant-related counting problem is considered using computer vision technologies under unconstrained field-based environment. With 361 field images collected in four experimental fields across China between 2010 and 2015 and corresponding manually-labelled dotted annotations, a novel Maize Tassels Counting ( MTC ) dataset is created and will be released with this paper. To alleviate the in-field challenges, a deep convolutional neural network-based approach termed TasselNet is proposed. TasselNet can achieve good adaptability to in-field variations via modelling the local visual characteristics of field images and regressing the local counts of maize tassels. Extensive results on the MTC dataset demonstrate that TasselNet outperforms other state-of-the-art approaches by large margins and achieves the overall best counting

  19. TasselNet: counting maize tassels in the wild via local counts regression network

    Directory of Open Access Journals (Sweden)

    Hao Lu

    2017-11-01

    Full Text Available Abstract Background Accurately counting maize tassels is important for monitoring the growth status of maize plants. This tedious task, however, is still mainly done by manual efforts. In the context of modern plant phenotyping, automating this task is required to meet the need of large-scale analysis of genotype and phenotype. In recent years, computer vision technologies have experienced a significant breakthrough due to the emergence of large-scale datasets and increased computational resources. Naturally image-based approaches have also received much attention in plant-related studies. Yet a fact is that most image-based systems for plant phenotyping are deployed under controlled laboratory environment. When transferring the application scenario to unconstrained in-field conditions, intrinsic and extrinsic variations in the wild pose great challenges for accurate counting of maize tassels, which goes beyond the ability of conventional image processing techniques. This calls for further robust computer vision approaches to address in-field variations. Results This paper studies the in-field counting problem of maize tassels. To our knowledge, this is the first time that a plant-related counting problem is considered using computer vision technologies under unconstrained field-based environment. With 361 field images collected in four experimental fields across China between 2010 and 2015 and corresponding manually-labelled dotted annotations, a novel Maize Tassels Counting (MTC dataset is created and will be released with this paper. To alleviate the in-field challenges, a deep convolutional neural network-based approach termed TasselNet is proposed. TasselNet can achieve good adaptability to in-field variations via modelling the local visual characteristics of field images and regressing the local counts of maize tassels. Extensive results on the MTC dataset demonstrate that TasselNet outperforms other state-of-the-art approaches by large

  20. Experimental evaluation of the extended Dytlewski-style dead time correction formalism for neutron multiplicity counting

    Science.gov (United States)

    Lockhart, M.; Henzlova, D.; Croft, S.; Cutler, T.; Favalli, A.; McGahee, Ch.; Parker, R.

    2018-01-01

    Over the past few decades, neutron multiplicity counting has played an integral role in Special Nuclear Material (SNM) characterization pertaining to nuclear safeguards. Current neutron multiplicity analysis techniques use singles, doubles, and triples count rates because a methodology to extract and dead time correct higher order count rates (i.e. quads and pents) was not fully developed. This limitation is overcome by the recent extension of a popular dead time correction method developed by Dytlewski. This extended dead time correction algorithm, named Dytlewski-Croft-Favalli(DCF), is detailed in reference Croft and Favalli (2017), which gives an extensive explanation of the theory and implications of this new development. Dead time corrected results can then be used to assay SNM by inverting a set of extended point model equations which as well have only recently been formulated. The current paper discusses and presents the experimental evaluation of practical feasibility of the DCF dead time correction algorithm to demonstrate its performance and applicability in nuclear safeguards applications. In order to test the validity and effectiveness of the dead time correction for quads and pents, 252Cf and SNM sources were measured in high efficiency neutron multiplicity counters at the Los Alamos National Laboratory (LANL) and the count rates were extracted up to the fifth order and corrected for dead time. In order to assess the DCF dead time correction, the corrected data is compared to traditional dead time correction treatment within INCC. The DCF dead time correction is found to provide adequate dead time treatment for broad range of count rates available in practical applications.

  1. Isospectral discrete and quantum graphs with the same flip counts and nodal counts

    Science.gov (United States)

    Juul, Jonas S.; Joyner, Christopher H.

    2018-06-01

    The existence of non-isomorphic graphs which share the same Laplace spectrum (to be referred to as isospectral graphs) leads naturally to the following question: what additional information is required in order to resolve isospectral graphs? It was suggested by Band, Shapira and Smilansky that this might be achieved by either counting the number of nodal domains or the number of times the eigenfunctions change sign (the so-called flip count) (Band et al 2006 J. Phys. A: Math. Gen. 39 13999–4014 Band and Smilansky 2007 Eur. Phys. J. Spec. Top. 145 171–9). Recent examples of (discrete) isospectral graphs with the same flip count and nodal count have been constructed by Ammann by utilising Godsil–McKay switching (Ammann private communication). Here, we provide a simple alternative mechanism that produces systematic examples of both discrete and quantum isospectral graphs with the same flip and nodal counts.

  2. Simple analytical technique for liquid scintillation counting of environmental carbon-14 using gel suspension method

    International Nuclear Information System (INIS)

    Okai, Tomio; Wakabayashi, Genichiro; Nagao, Kenjiro; Matoba, Masaru; Ohura, Hirotaka; Momoshima, Noriyuki; Kawamura, Hidehisa

    2000-01-01

    A simple analytical technique for liquid scintillation counting of environmental 14 C was developed. Commercially available gelling agent, N-lauroyl-L -glutamic -α,γ-dibutylamide, was used for the gel-formation of the samples (gel suspension method) and for the subsequent liquid scintillation counting of 14 C in the form of CaCO 3 . Our procedure for sample preparation is much simpler than that of the conventional methods and requires no special equipment. Self absorption, stability and reproducibility of gel suspension samples were investigated in order to evaluate the characteristics of the gel suspension method for 14 C activity measurement. The self absorption factor is about 70% and slightly decrease as CaCO 3 weight increase. This is considered to be mainly due to the absorption of β-rays and scintillation light by the CaCO 3 sample itself. No change of the counting rate for the gel suspension sample was observed for more than 2 years after the sample preparation. Four samples were used for checking the reproducibility of the sample preparation method. The same values were obtained for the counting rate of 24 C activity within the counting error. No change of the counting rate was observed for the 're-gelated' sample. These results show that the gel suspension method is appropriate for the 14 C activity measurement by the liquid scintillation counting method and useful for a long-term preservation of the sample for repeated measurement. The above analytical technique was applied to actual environmental samples in Fukuoka prefecture, Japan. Results obtained were comparable with those by other researchers and appear to be reasonable. Therefore, the newly developed technique is useful for the routine monitoring of environmental 14 C. (author)

  3. High energy resolution and high count rate gamma spectrometry measurement of primary coolant of generation 4 sodium-cooled fast reactor

    International Nuclear Information System (INIS)

    Coulon, R.

    2010-01-01

    Sodium-cooled Fast Reactors are under development for the fourth generation of nuclear reactor. Breeders reactors could gives solutions for the need of energy and the preservation of uranium resources. An other purpose is the radioactive wastes production reduction by transmutation and the control of non-proliferation using a closed-cycle. These thesis shows safety and profit advantages that could be obtained by a new generation of gamma spectrometry system for SFR. Now, the high count rate abilities, allow us to study new methods of accurate power measurement and fast clad failure detection. Simulations have been done and an experimental test has been performed at the French Phenix SFR of the CEA Marcoule showing promising results for these new measurements. (author) [fr

  4. Counting It Twice.

    Science.gov (United States)

    Schattschneider, Doris

    1991-01-01

    Provided are examples from many domains of mathematics that illustrate the Fubini Principle in its discrete version: the value of a summation over a rectangular array is independent of the order of summation. Included are: counting using partitions as in proof by pictures, combinatorial arguments, indirect counting as in the inclusion-exclusion…

  5. Low White Blood Cell Count

    Science.gov (United States)

    Symptoms Low white blood cell count By Mayo Clinic Staff A low white blood cell count (leukopenia) is a decrease ... of white blood cell (neutrophil). The definition of low white blood cell count varies from one medical ...

  6. Temporal aggregation of migration counts can improve accuracy and precision of trends

    Directory of Open Access Journals (Sweden)

    Tara L. Crewe

    2016-12-01

    Full Text Available Temporal replicate counts are often aggregated to improve model fit by reducing zero-inflation and count variability, and in the case of migration counts collected hourly throughout a migration, allows one to ignore nonindependence. However, aggregation can represent a loss of potentially useful information on the hourly or seasonal distribution of counts, which might impact our ability to estimate reliable trends. We simulated 20-year hourly raptor migration count datasets with known rate of change to test the effect of aggregating hourly counts to daily or annual totals on our ability to recover known trend. We simulated data for three types of species, to test whether results varied with species abundance or migration strategy: a commonly detected species, e.g., Northern Harrier, Circus cyaneus; a rarely detected species, e.g., Peregrine Falcon, Falco peregrinus; and a species typically counted in large aggregations with overdispersed counts, e.g., Broad-winged Hawk, Buteo platypterus. We compared accuracy and precision of estimated trends across species and count types (hourly/daily/annual using hierarchical models that assumed a Poisson, negative binomial (NB or zero-inflated negative binomial (ZINB count distribution. We found little benefit of modeling zero-inflation or of modeling the hourly distribution of migration counts. For the rare species, trends analyzed using daily totals and an NB or ZINB data distribution resulted in a higher probability of detecting an accurate and precise trend. In contrast, trends of the common and overdispersed species benefited from aggregation to annual totals, and for the overdispersed species in particular, trends estimating using annual totals were more precise, and resulted in lower probabilities of estimating a trend (1 in the wrong direction, or (2 with credible intervals that excluded the true trend, as compared with hourly and daily counts.

  7. Estimated average annual rate of change of CD4(+) T-cell counts in patients on combination antiretroviral therapy

    DEFF Research Database (Denmark)

    Mocroft, Amanda; Phillips, Andrew N; Ledergerber, Bruno

    2010-01-01

    BACKGROUND: Patients receiving combination antiretroviral therapy (cART) might continue treatment with a virologically failing regimen. We sought to identify annual change in CD4(+) T-cell count according to levels of viraemia in patients on cART. METHODS: A total of 111,371 CD4(+) T-cell counts...... and viral load measurements in 8,227 patients were analysed. Annual change in CD4(+) T-cell numbers was estimated using mixed models. RESULTS: After adjustment, the estimated average annual change in CD4(+) T-cell count significantly increased when viral load was cells/mm(3), 95......% confidence interval [CI] 26.6-34.3), was stable when viral load was 500-9,999 copies/ml (3.1 cells/mm(3), 95% CI -5.3-11.5) and decreased when viral load was >/=10,000 copies/ml (-14.8 cells/mm(3), 95% CI -4.5--25.1). Patients taking a boosted protease inhibitor (PI) regimen had more positive annual CD4(+) T-cell...

  8. Platelet Counts in Insoluble Platelet-Rich Fibrin Clots: A Direct Method for Accurate Determination

    Directory of Open Access Journals (Sweden)

    Yutaka Kitamura

    2018-02-01

    Full Text Available Platelet-rich fibrin (PRF clots have been used in regenerative dentistry most often, with the assumption that growth factor levels are concentrated in proportion to the platelet concentration. Platelet counts in PRF are generally determined indirectly by platelet counting in other liquid fractions. This study shows a method for direct estimation of platelet counts in PRF. To validate this method by determination of the recovery rate, whole-blood samples were obtained with an anticoagulant from healthy donors, and platelet-rich plasma (PRP fractions were clotted with CaCl2 by centrifugation and digested with tissue-plasminogen activator. Platelet counts were estimated before clotting and after digestion using an automatic hemocytometer. The method was then tested on PRF clots. The quality of platelets was examined by scanning electron microscopy and flow cytometry. In PRP-derived fibrin matrices, the recovery rate of platelets and white blood cells was 91.6 and 74.6%, respectively, after 24 h of digestion. In PRF clots associated with small and large red thrombi, platelet counts were 92.6 and 67.2% of the respective total platelet counts. These findings suggest that our direct method is sufficient for estimating the number of platelets trapped in an insoluble fibrin matrix and for determining that platelets are distributed in PRF clots and red thrombi roughly in proportion to their individual volumes. Therefore, we propose this direct digestion method for more accurate estimation of platelet counts in most types of platelet-enriched fibrin matrix.

  9. Petroleum production at Maximum Efficient Rate Naval Petroleum Reserve No. 1 (Elk Hills), Kern County, California

    International Nuclear Information System (INIS)

    1993-07-01

    This document provides an analysis of the potential impacts associated with the proposed action, which is continued operation of Naval Petroleum Reserve No. I (NPR-1) at the Maximum Efficient Rate (MER) as authorized by Public law 94-258, the Naval Petroleum Reserves Production Act of 1976 (Act). The document also provides a similar analysis of alternatives to the proposed action, which also involve continued operations, but under lower development scenarios and lower rates of production. NPR-1 is a large oil and gas field jointly owned and operated by the federal government and Chevron U.SA Inc. (CUSA) pursuant to a Unit Plan Contract that became effective in 1944; the government's interest is approximately 78% and CUSA's interest is approximately 22%. The government's interest is under the jurisdiction of the United States Department of Energy (DOE). The facility is approximately 17,409 acres (74 square miles), and it is located in Kern County, California, about 25 miles southwest of Bakersfield and 100 miles north of Los Angeles in the south central portion of the state. The environmental analysis presented herein is a supplement to the NPR-1 Final Environmental Impact Statement of that was issued by DOE in 1979 (1979 EIS). As such, this document is a Supplemental Environmental Impact Statement (SEIS)

  10. Partial-Interval Estimation of Count: Uncorrected and Poisson-Corrected Error Levels

    Science.gov (United States)

    Yoder, Paul J.; Ledford, Jennifer R.; Harbison, Amy L.; Tapp, Jon T.

    2018-01-01

    A simulation study that used 3,000 computer-generated event streams with known behavior rates, interval durations, and session durations was conducted to test whether the main and interaction effects of true rate and interval duration affect the error level of uncorrected and Poisson-transformed (i.e., "corrected") count as estimated by…

  11. Long term observation on absolute lymphocyte counts in the adult health study sample, Hiroshima and Nagasaki

    International Nuclear Information System (INIS)

    Oesterle, S.N.; Norman, J.E. Jr.

    1980-01-01

    Total peripheral blood lymphocytes were evaluated by age and exposure status in the Adult Health Study population during three examination cycles between 1958 and 1972. No radiation effect was observed, but a significant drop in the absolute lymphocyte counts of those aged 70 years and over and a corresponding maximum for persons aged 50 - 59 was observed. (author)

  12. Liquid scintillation counting analysis of cadmium-109

    International Nuclear Information System (INIS)

    Robinson, M.K.; Barfuss, D.W.

    1991-01-01

    Recently the authors have used radiolabled cadmium-109 to measure the transport of inorganic cadmium in renal proximal tubules. An anomaly discovered in the liquid scintillation counting analysis of Cd-109 which is not attributable to normal decay; it consists of a significant decrease in the measured count rate of small amounts of sample. The objective is to determine whether the buffer solution used in the membrane transport studies is causing precipitation of the cadmium or whether cadmium is being adsorbed by the glass. It was important to determine whether the procedure could be modified to correct this problem. The problem does not appear to be related to the use of the buffer or to adsorption of Cd onto glass. Correction based on using triated L-glucose in all of these experiments and calculating a correction factor for the concentration of cadmium

  13. Evaluation of high-energy electron detectors for probing the inner magnetosphere under high-counting condition

    International Nuclear Information System (INIS)

    Tamada, Yukihiro; Takashima, Takeshi; Mitani, Takefumi; Miyake, Wataru

    2013-01-01

    An ERG (Energization and Radiation in Geospace) satellite will be launched to study the acceleration processes of energetic particles in the radiation belt surrounding the earth. It is very important to reveal the acceleration process of high-energy particles for both science and the application to space weather forecast. Drastic increases of high-energy electrons in the radiation belt is sometimes observed during a geomagnetic storm. When a large magnetic storm occurs, energetic electron count rates may exceed flux limits expected in the nominal design and large number of incident electrons leading to detection loss. The purpose of this study is to demonstrate that the count rate range of a single detection on board ERG satellite can be expanded by means of reading circuit operations to decrease an area of detection. In our ground experiment, we also found an unexpected result that count peaks shift to the higher energy side under high counting conditions. (author)

  14. Characterization of spectrometric photon-counting X-ray detectors at different pitches

    International Nuclear Information System (INIS)

    Jurdit, M.; Moulin, V.; Ouvrier-Buffet, P.; Verger, L.; Brambilla, A.; Radisson, P.

    2017-01-01

    There is growing interest in energy-sensitive photon-counting detectors based on high flux X-ray imaging. Their potential applications include medical imaging, non-destructive testing and security. Innovative detectors of this type will need to count individual photons and sort them into selected energy bins, at several million counts per second and per mm 2 . Cd(Zn)Te detector grade materials with a thickness of 1.5 to 3 mm and pitches from 800 μm down to 200 μm were assembled onto interposer boards. These devices were tested using in-house-developed full-digital fast readout electronics. The 16-channel demonstrators, with 256 energy bins, were experimentally characterized by determining spectral resolution, count rate, and charge sharing, which becomes challenging at low pitch. Charge sharing correction was found to efficiently correct X-ray spectra up to 40 × 10 6 incident photons.s −1 .mm −2 .

  15. Novel fast-neutron activation counter for high repetition rate measurements

    International Nuclear Information System (INIS)

    Mahmood, S.; Springham, S. V.; Zhang, T.; Rawat, R. S.; Tan, T. L.; Krishnan, M.; Beg, F. N.; Lee, S.; Schmidt, H.; Lee, P.

    2006-01-01

    A fast-neutron beryllium activation counter has been constructed for neutron measurements on a high repetition rate deuterium plasma focus. Beryllium activation is especially suitable for measurements of DD neutron yields. The cross section for the relevant reaction, 9 Be(n,α) 6 He, results in a maximum sensitivity at the characteristic energy of the DD neutrons (∼2.5 MeV) and practically no sensitivity to neutrons with energies 6 He enabled the shot-to-shot neutron yield from the plasma focus to be measured for repetition rates from 0.2 to 3 Hz (and for a range of deuterium gas pressures). With careful analysis, the shot-to-shot yield can be measured up to a maximum repetition rate of 3 Hz, beyond which the pileup of counts from the previous shots reduces the accuracy of the measurements to an unacceptable level. This new beryllium activation counter has been cross-checked against an indium activation counter to obtain absolute neutron yields. At a charging voltage of 12.5 kV (bank energy of 2.2 kJ), the average neutron yield was found to be (7.9±0.7)x10 7 per shot (standard deviation of 4x10 7 ). It was found that activation of the plasma focus construction materials (especially aluminum) must be taken into account

  16. High Flux Energy-Resolved Photon-Counting X-Ray Imaging Arrays with CdTe and CdZnTe for Clinical CT

    International Nuclear Information System (INIS)

    Barber, William C.; Hartsough, Neal E.; Gandhi, Thulasidharan; Iwanczyk, Jan S.; Wessel, Jan C.; Nygard, Einar; Malakhov, Nail; Wawrzyniak, Gregor; Dorholt, Ole; Danielsen, Roar

    2013-06-01

    We have fabricated fast room-temperature energy dispersive photon counting x-ray imaging arrays using pixellated cadmium zinc (CdTe) and cadmium zinc telluride (CdZnTe) semiconductors. We have also fabricated fast application specific integrated circuits (ASICs) with a two dimensional (2D) array of inputs for readout from the CdZnTe sensors. The new CdTe and CdZnTe sensors have a 2D array of pixels with a 0.5 mm pitch and can be tiled in 2D. The new 2D ASICs have four energy discriminators per pixel with a linear energy response across the entire dynamic range for clinical CT. The ASICs can also be tiled in 2D and are designed to fit within the active area of the 2D sensors. We have measured several important performance parameters including; an output count rate (OCR) in excess of 20 million counts per second per square mm, an energy resolution of 7 keV full width at half maximum (FWHM) across the entire dynamic range, and a noise floor less than 20 keV. This is achieved by directly interconnecting the ASIC inputs to the pixels of the CdTE and CdZnTe sensors incurring very little additional capacitance. We present a comparison of the performance of the CdTe and CdZnTe sensors including the OCR, FWHM energy resolution, and noise floor. (authors)

  17. Virtual point detector: On the interpolation and extrapolation of scintillation detectors counting efficiencies

    International Nuclear Information System (INIS)

    Presler, Oren; German, Uzi; Pushkarsky, Vitaly; Alfassi, Zeev B.

    2006-01-01

    The concept of transforming the detector volume to a virtual point detector, in order to facilitate efficiency evaluations for different source locations, was proposed in the past for HPGe and Ge(Li) detectors. The validity of this model for NaI(Tl) and BGO scintillation detectors was studied in the present work. It was found that for both scintillation detectors, the point detector model does not seem to fit too well to the experimental data, for the whole range of source-to-detector distances; however, for source-to-detector cap distances larger than 4 cm, the accuracy was found to be high. A two-parameter polynomial expression describing the dependence of the normalized count rate versus the source-to-detector distance was fitted to the experimental data. For this fit, the maximum deviations are up to about 12%. These deviations are much smaller than the values obtained by applying the virtual point concept, even for distances greater than 4 cm, thus the polynomial fitting is to be preferred for scintillation detectors

  18. The real-time fitting of radioactive decay curves. Pt. 3. Counting during sampling

    International Nuclear Information System (INIS)

    Hartley, B.M.

    1994-01-01

    An analysis of a least-squares method for the real-time fitting of the theoretical total count function to the actual total count from radioactive decays has been given previously for the case where counting takes place after a sample is taken. The counting may be done in a number of different counting systems which distinguish between different types or energies of radiation emitted from the sample. The method would allow real-time determination of the numbers of atoms and hence activities of the individual isotopes present and has been designated the Time Evolved Least-Squares method (TELS). If the radioactivity which is to be measured exists as an aerosol or in a form where a sample is taken at a constant rate it may be possible to count during sampling and by so doing reduce the total time required to determine the activity of the individual isotopes present. The TELS method is extended here to the case where counting and the evaluation of the activity takes place concurrently with the sampling. The functions which need to be evaluated are derived and the calculations required to implement the method are discussed. As with the TELS method of counting after sampling the technique of counting during sampling and the simultaneous evaluation of activity could be achieved in real-time. Results of testing the method by computer simulation for two counting schemes for the descendants of radon are presented. ((orig.))

  19. A transimpedance CMOS multichannel amplifier with a 50 Ω-wide output range buffer for high counting rate applications

    International Nuclear Information System (INIS)

    Haralabidis, N.; Loukas, D.; Misiakos, K.; Katsafouros, S.

    1997-01-01

    A fast transimpedance multichannel amplifier has been designed, fabricated in CMOS 1.2-microm technology and tested. Each channel consists of a current sensitive preamplifier followed by a voltage amplification stage and an on-chip buffer able to drive 50 Ω loads with an output range of ±800 mV. Measured peaking time at the output is 40 ns and the circuit recovers to baseline in 90 ns. This results in a counting capability of more than 10 7 hits/s. Signals of both polarities can be handled. The first two stages consume a total of 2 mW per channel and the 50 Ω buffer consumes another 17 mW. The equivalent noise charge (ENC) is 1,100 e - rms with a slope of 40e - /pF. The IC is intended for use in gas and solid-state detectors with high particle rate and extensive charge release as in high energy calorimetry

  20. Effects of lek count protocols on greater sage-grouse population trend estimates

    Science.gov (United States)

    Monroe, Adrian; Edmunds, David; Aldridge, Cameron L.

    2016-01-01

    Annual counts of males displaying at lek sites are an important tool for monitoring greater sage-grouse populations (Centrocercus urophasianus), but seasonal and diurnal variation in lek attendance may increase variance and bias of trend analyses. Recommendations for protocols to reduce observation error have called for restricting lek counts to within 30 minutes of sunrise, but this may limit the number of lek counts available for analysis, particularly from years before monitoring was widely standardized. Reducing the temporal window for conducting lek counts also may constrain the ability of agencies to monitor leks efficiently. We used lek count data collected across Wyoming during 1995−2014 to investigate the effect of lek counts conducted between 30 minutes before and 30, 60, or 90 minutes after sunrise on population trend estimates. We also evaluated trends across scales relevant to management, including statewide, within Working Group Areas and Core Areas, and for individual leks. To further evaluate accuracy and precision of trend estimates from lek count protocols, we used simulations based on a lek attendance model and compared simulated and estimated values of annual rate of change in population size (λ) from scenarios of varying numbers of leks, lek count timing, and count frequency (counts/lek/year). We found that restricting analyses to counts conducted within 30 minutes of sunrise generally did not improve precision of population trend estimates, although differences among timings increased as the number of leks and count frequency decreased. Lek attendance declined >30 minutes after sunrise, but simulations indicated that including lek counts conducted up to 90 minutes after sunrise can increase the number of leks monitored compared to trend estimates based on counts conducted within 30 minutes of sunrise. This increase in leks monitored resulted in greater precision of estimates without reducing accuracy. Increasing count

  1. Hanford whole body counting manual

    International Nuclear Information System (INIS)

    Palmer, H.E.; Brim, C.P.; Rieksts, G.A.; Rhoads, M.C.

    1987-05-01

    This document, a reprint of the Whole Body Counting Manual, was compiled to train personnel, document operation procedures, and outline quality assurance procedures. The current manual contains information on: the location, availability, and scope of services of Hanford's whole body counting facilities; the administrative aspect of the whole body counting operation; Hanford's whole body counting facilities; the step-by-step procedure involved in the different types of in vivo measurements; the detectors, preamplifiers and amplifiers, and spectroscopy equipment; the quality assurance aspect of equipment calibration and recordkeeping; data processing, record storage, results verification, report preparation, count summaries, and unit cost accounting; and the topics of minimum detectable amount and measurement accuracy and precision. 12 refs., 13 tabs

  2. A method for the measurement of the intrinsic dead time of a counting system

    International Nuclear Information System (INIS)

    Wyllie, H.A.

    1989-01-01

    Equations are derived for (a) the determination of the intrinsic dead time of a counting system in the components preceding the paralysis unit which imposes the set dead time, and (b) a more accurate correction of count rates in a single-channel system, taking into account the extension of the set dead time by the intrinsic dead time. (author)

  3. A high dynamic range pulse counting detection system for mass spectrometry.

    Science.gov (United States)

    Collings, Bruce A; Dima, Martian D; Ivosev, Gordana; Zhong, Feng

    2014-01-30

    A high dynamic range pulse counting system has been developed that demonstrates an ability to operate at up to 2e8 counts per second (cps) on a triple quadrupole mass spectrometer. Previous pulse counting detection systems have typically been limited to about 1e7 cps at the upper end of the systems dynamic range. Modifications to the detection electronics and dead time correction algorithm are described in this paper. A high gain transimpedance amplifier is employed that allows a multi-channel electron multiplier to be operated at a significantly lower bias potential than in previous pulse counting systems. The system utilises a high-energy conversion dynode, a multi-channel electron multiplier, a high gain transimpedance amplifier, non-paralysing detection electronics and a modified dead time correction algorithm. Modification of the dead time correction algorithm is necessary due to a characteristic of the pulse counting electronics. A pulse counting detection system with the capability to count at ion arrival rates of up to 2e8 cps is described. This is shown to provide a linear dynamic range of nearly five orders of magnitude for a sample of aprazolam with concentrations ranging from 0.0006970 ng/mL to 3333 ng/mL while monitoring the m/z 309.1 → m/z 205.2 transition. This represents an upward extension of the detector's linear dynamic range of about two orders of magnitude. A new high dynamic range pulse counting system has been developed demonstrating the ability to operate at up to 2e8 cps on a triple quadrupole mass spectrometer. This provides an upward extension of the detector's linear dynamic range by about two orders of magnitude over previous pulse counting systems. Copyright © 2013 John Wiley & Sons, Ltd.

  4. CalCOFI Egg Counts

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Fish egg counts and standardized counts for eggs captured in CalCOFI icthyoplankton nets (primarily vertical [Calvet or Pairovet], oblique [bongo or ring nets], and...

  5. Count-doubling time safety circuit

    International Nuclear Information System (INIS)

    Keefe, D.J.; McDowell, W.P.; Rusch, G.K.

    1981-01-01

    There is provided a nuclear reactor count-factor-increase time monitoring circuit which includes a pulse-type neutron detector, and means for counting the number of detected pulses during specific time periods. Counts are compared and the comparison is utilized to develop a reactor scram signal, if necessary

  6. Count-doubling time safety circuit

    Science.gov (United States)

    Rusch, Gordon K.; Keefe, Donald J.; McDowell, William P.

    1981-01-01

    There is provided a nuclear reactor count-factor-increase time monitoring circuit which includes a pulse-type neutron detector, and means for counting the number of detected pulses during specific time periods. Counts are compared and the comparison is utilized to develop a reactor scram signal, if necessary.

  7. The Big Pumpkin Count.

    Science.gov (United States)

    Coplestone-Loomis, Lenny

    1981-01-01

    Pumpkin seeds are counted after students convert pumpkins to jack-o-lanterns. Among the activities involved, pupils learn to count by 10s, make estimates, and to construct a visual representation of 1,000. (MP)

  8. Galaxy number counts: Pt. 2

    International Nuclear Information System (INIS)

    Metcalfe, N.; Shanks, T.; Fong, R.; Jones, L.R.

    1991-01-01

    Using the Prime Focus CCD Camera at the Isaac Newton Telescope we have determined the form of the B and R galaxy number-magnitude count relations in 12 independent fields for 21 m ccd m and 19 m ccd m 5. The average galaxy count relations lie in the middle of the wide range previously encompassed by photographic data. The field-to-field variation of the counts is small enough to define the faint (B m 5) galaxy count to ±10 per cent and this variation is consistent with that expected from galaxy clustering considerations. Our new data confirm that the B, and also the R, galaxy counts show evidence for strong galaxy luminosity evolution, and that the majority of the evolving galaxies are of moderately blue colour. (author)

  9. SU-G-IeP4-12: Performance of In-111 Coincident Gamma-Ray Counting: A Monte Carlo Simulation

    Energy Technology Data Exchange (ETDEWEB)

    Pahlka, R; Kappadath, S; Mawlawi, O [UT MD Anderson Cancer Center, Houston, TX (United States)

    2016-06-15

    Purpose: The decay of In-111 results in a non-isotropic gamma-ray cascade, which is normally imaged using a gamma camera. Creating images with a gamma camera using coincident gamma-rays from In-111 has not been previously studied. Our objective was to explore the feasibility of imaging this cascade as coincidence events and to determine the optimal timing resolution and source activity using Monte Carlo simulations. Methods: GEANT4 was used to simulate the decay of the In-111 nucleus and to model the gamma camera. Each photon emission was assigned a timestamp, and the time delay and angular separation for the second gamma-ray in the cascade was consistent with the known intermediate state half-life of 85ns. The gamma-rays are transported through a model of a Siemens dual head Symbia “S” gamma camera with a 5/8-inch thick crystal and medium energy collimators. A true coincident event was defined as a single 171keV gamma-ray followed by a single 245keV gamma-ray within a specified time window (or vice versa). Several source activities (ranging from 10uCi to 5mCi) with and without incorporation of background counts were then simulated. Each simulation was analyzed using varying time windows to assess random events. The noise equivalent count rate (NECR) was computed based on the number of true and random counts for each combination of activity and time window. No scatter events were assumed since sources were simulated in air. Results: As expected, increasing the timing window increased the total number of observed coincidences albeit at the expense of true coincidences. A timing window range of 200–500ns maximizes the NECR at clinically-used source activities. The background rate did not significantly alter the maximum NECR. Conclusion: This work suggests coincident measurements of In-111 gamma-ray decay can be performed with commercial gamma cameras at clinically-relevant activities. Work is ongoing to assess useful clinical applications.

  10. Digital coincidence counting

    Science.gov (United States)

    Buckman, S. M.; Ius, D.

    1996-02-01

    This paper reports on the development of a digital coincidence-counting system which comprises a custom-built data acquisition card and associated PC software. The system has been designed to digitise the pulse-trains from two radiation detectors at a rate of 20 MSamples/s with 12-bit resolution. Through hardware compression of the data, the system can continuously record both individual pulse-shapes and the time intervals between pulses. Software-based circuits are used to process the stored pulse trains. These circuits are constructed simply by linking together icons representing various components such as coincidence mixers, time delays, single-channel analysers, deadtimes and scalers. This system enables a pair of pulse trains to be processed repeatedly using any number of different methods. Some preliminary results are presented in order to demonstrate the versatility and efficiency of this new method.

  11. Digital coincidence counting

    International Nuclear Information System (INIS)

    Buckman, S.M.; Ius, D.

    1996-01-01

    This paper reports on the development of a digital coincidence-counting system which comprises a custom-built data acquisition card and associated PC software. The system has been designed to digitise the pulse-trains from two radiation detectors at a rate of 20 MSamples/s with 12-bit resolution. Through hardware compression of the data, the system can continuously record both individual pulse-shapes and the time intervals between pulses. Software-based circuits are used to process the stored pulse trains. These circuits are constructed simply by linking together icons representing various components such as coincidence mixers, time delays, single-channel analysers, deadtimes and scalers. This system enables a pair of pulse trains to be processed repeatedly using any number of different methods. Some preliminary results are presented in order to demonstrate the versatility and efficiency of this new method. (orig.)

  12. Counting statistics in radioactivity measurements

    International Nuclear Information System (INIS)

    Martin, J.

    1975-01-01

    The application of statistical methods to radioactivity measurement problems is analyzed in several chapters devoted successively to: the statistical nature of radioactivity counts; the application to radioactive counting of two theoretical probability distributions, Poisson's distribution law and the Laplace-Gauss law; true counting laws; corrections related to the nature of the apparatus; statistical techniques in gamma spectrometry [fr

  13. Should measurement of maximum urinary flow rate and residual urine volume be a part of a "minimal care" assessment programme in female incontinence?

    DEFF Research Database (Denmark)

    Sander, Pia; Mouritsen, L; Andersen, J Thorup

    2002-01-01

    OBJECTIVE: The aim of this study was to evaluate the value of routine measurements of urinary flow rate and residual urine volume as a part of a "minimal care" assessment programme for women with urinary incontinence in detecting clinical significant bladder emptying problems. MATERIAL AND METHODS....... Twenty-six per cent had a maximum flow rate less than 15 ml/s, but only 4% at a voided volume > or =200 ml. Residual urine more than 149 ml was found in 6%. Two women had chronic retention with overflow incontinence. Both had typical symptoms with continuous leakage, stranguria and chronic cystitis...

  14. Correlation between total lymphocyte count, hemoglobin, hematocrit and CD4 count in HIV patients in Nigeria.

    Science.gov (United States)

    Emuchay, Charles Iheanyichi; Okeniyi, Shemaiah Olufemi; Okeniyi, Joshua Olusegun

    2014-04-01

    The expensive and technology limited setting of CD4 count testing is a major setback to the initiation of HAART in a resource limited country like Nigeria. Simple and inexpensive tools such as Hemoglobin (Hb) measurement and Total Lymphocyte Count (TLC) are recommended as substitute marker. In order to assess the correlations of these parameters with CD4 count, 100 "apparently healthy" male volunteers tested HIV positive aged ≥ 20 years but ≤ 40 years were recruited and from whom Hb, Hct, TLC and CD4 count were obtained. The correlation coefficients, R, the Nash-Sutcliffe Coefficient of Efficiency (CoE) and the p-values of the ANOVA model of Hb, Hct and TLC with CD4 count were assessed. The assessments show that there is no significant relationship of any of these parameters with CD4 count and the correlation coefficients are very weak. This study shows that Hb, Hct and TLC cannot be substitute for CD4 count as this might lead to certain individuals' deprivation of required treatment.

  15. It counts who counts: an experimental evaluation of the importance of observer effects on spotlight count estimates

    DEFF Research Database (Denmark)

    Sunde, Peter; Jessen, Lonnie

    2013-01-01

    observers with respect to their ability to detect and estimate distance to realistic animal silhouettes at different distances. Detection probabilities were higher for observers experienced in spotlighting mammals than for inexperienced observers, higher for observers with a hunting background compared...... with non-hunters and decreased as function of age but were independent of sex or educational background. If observer-specific detection probabilities were applied to real counting routes, point count estimates from inexperienced observers without a hunting background would only be 43 % (95 % CI, 39...

  16. Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging.

    Science.gov (United States)

    Iwanczyk, Jan S; Nygård, Einar; Meirav, Oded; Arenson, Jerry; Barber, William C; Hartsough, Neal E; Malakhov, Nail; Wessel, Jan C

    2009-01-01

    The development of an innovative detector technology for photon-counting in X-ray imaging is reported. This new generation of detectors, based on pixellated cadmium telluride (CdTe) and cadmium zinc telluride (CZT) detector arrays electrically connected to application specific integrated circuits (ASICs) for readout, will produce fast and highly efficient photon-counting and energy-dispersive X-ray imaging. There are a number of applications that can greatly benefit from these novel imagers including mammography, planar radiography, and computed tomography (CT). Systems based on this new detector technology can provide compositional analysis of tissue through spectroscopic X-ray imaging, significantly improve overall image quality, and may significantly reduce X-ray dose to the patient. A very high X-ray flux is utilized in many of these applications. For example, CT scanners can produce ~100 Mphotons/mm(2)/s in the unattenuated beam. High flux is required in order to collect sufficient photon statistics in the measurement of the transmitted flux (attenuated beam) during the very short time frame of a CT scan. This high count rate combined with a need for high detection efficiency requires the development of detector structures that can provide a response signal much faster than the transit time of carriers over the whole detector thickness. We have developed CdTe and CZT detector array structures which are 3 mm thick with 16×16 pixels and a 1 mm pixel pitch. These structures, in the two different implementations presented here, utilize either a small pixel effect or a drift phenomenon. An energy resolution of 4.75% at 122 keV has been obtained with a 30 ns peaking time using discrete electronics and a (57)Co source. An output rate of 6×10(6) counts per second per individual pixel has been obtained with our ASIC readout electronics and a clinical CT X-ray tube. Additionally, the first clinical CT images, taken with several of our prototype photon-counting and

  17. You can count on the motor cortex: Finger counting habits modulate motor cortex activation evoked by numbers

    Science.gov (United States)

    Tschentscher, Nadja; Hauk, Olaf; Fischer, Martin H.; Pulvermüller, Friedemann

    2012-01-01

    The embodied cognition framework suggests that neural systems for perception and action are engaged during higher cognitive processes. In an event-related fMRI study, we tested this claim for the abstract domain of numerical symbol processing: is the human cortical motor system part of the representation of numbers, and is organization of numerical knowledge influenced by individual finger counting habits? Developmental studies suggest a link between numerals and finger counting habits due to the acquisition of numerical skills through finger counting in childhood. In the present study, digits 1 to 9 and the corresponding number words were presented visually to adults with different finger counting habits, i.e. left- and right-starters who reported that they usually start counting small numbers with their left and right hand, respectively. Despite the absence of overt hand movements, the hemisphere contralateral to the hand used for counting small numbers was activated when small numbers were presented. The correspondence between finger counting habits and hemispheric motor activation is consistent with an intrinsic functional link between finger counting and number processing. PMID:22133748

  18. The Relationship between Pollen Count Levels and Prevalence of Japanese Cedar Pollinosis in Northeast Japan

    Directory of Open Access Journals (Sweden)

    Kohei Honda

    2013-01-01

    Conclusions: These results suggest that pollen count levels may correlate with the rate of sensitization for JC pollinosis, but may not affect the rate of onset among sensitized children in northeast Japan.

  19. Regularization parameter selection methods for ill-posed Poisson maximum likelihood estimation

    International Nuclear Information System (INIS)

    Bardsley, Johnathan M; Goldes, John

    2009-01-01

    In image processing applications, image intensity is often measured via the counting of incident photons emitted by the object of interest. In such cases, image data noise is accurately modeled by a Poisson distribution. This motivates the use of Poisson maximum likelihood estimation for image reconstruction. However, when the underlying model equation is ill-posed, regularization is needed. Regularized Poisson likelihood estimation has been studied extensively by the authors, though a problem of high importance remains: the choice of the regularization parameter. We will present three statistically motivated methods for choosing the regularization parameter, and numerical examples will be presented to illustrate their effectiveness

  20. Total lymphocyte count and subpopulation lymphocyte counts in relation to dietary intake and nutritional status of peritoneal dialysis patients.

    Science.gov (United States)

    Grzegorzewska, Alicja E; Leander, Magdalena

    2005-01-01

    Dietary deficiency causes abnormalities in circulating lymphocyte counts. For the present paper, we evaluated correlations between total and subpopulation lymphocyte counts (TLC, SLCs) and parameters of nutrition in peritoneal dialysis (PD) patients. Studies were carried out in 55 patients treated with PD for 22.2 +/- 11.4 months. Parameters of nutritional status included total body mass, lean body mass (LBM), body mass index (BMI), and laboratory indices [total protein, albumin, iron, ferritin, and total iron binding capacity (TIBC)]. The SLCs were evaluated using flow cytometry. Positive correlations were seen between TLC and dietary intake of niacin; TLC and CD8 and CD16+56 counts and energy delivered from protein; CD4 count and beta-carotene and monounsaturated fatty acids 17:1 intake; and CD19 count and potassium, copper, vitamin A, and beta-carotene intake. Anorexia negatively influenced CD19 count. Serum albumin showed correlations with CD4 and CD19 counts, and LBM with CD19 count. A higher CD19 count was connected with a higher red blood cell count, hemoglobin, and hematocrit. Correlations were observed between TIBC and TLC and CD3 and CD8 counts, and between serum Fe and TLC and CD3 and CD4 counts. Patients with a higher CD19 count showed a better clinical-laboratory score, especially less weakness. Patients with a higher CD4 count had less expressed insomnia. Quantities of ingested vitamins and minerals influence lymphocyte counts in the peripheral blood of PD patients. Evaluation of TLC and SLCs is helpful in monitoring the effectiveness of nutrition in these patients.

  1. Ultrafast photon counting applied to resonant scanning STED microscopy.

    Science.gov (United States)

    Wu, Xundong; Toro, Ligia; Stefani, Enrico; Wu, Yong

    2015-01-01

    To take full advantage of fast resonant scanning in super-resolution stimulated emission depletion (STED) microscopy, we have developed an ultrafast photon counting system based on a multigiga sample per second analogue-to-digital conversion chip that delivers an unprecedented 450 MHz pixel clock (2.2 ns pixel dwell time in each scan). The system achieves a large field of view (∼50 × 50 μm) with fast scanning that reduces photobleaching, and advances the time-gated continuous wave STED technology to the usage of resonant scanning with hardware-based time-gating. The assembled system provides superb signal-to-noise ratio and highly linear quantification of light that result in superior image quality. Also, the system design allows great flexibility in processing photon signals to further improve the dynamic range. In conclusion, we have constructed a frontier photon counting image acquisition system with ultrafast readout rate, excellent counting linearity, and with the capacity of realizing resonant-scanning continuous wave STED microscopy with online time-gated detection. © 2014 The Authors Journal of Microscopy © 2014 Royal Microscopical Society.

  2. Recent Advances in Digital Coincidence Counting for Radionuclide Metrology

    International Nuclear Information System (INIS)

    Keightley, John; Bobin, Christophe; Bouchard, Jacques; Capogni, Marco; Loreti, Stefano; Roteta, Miguel

    2013-06-01

    The radioactivity measurement techniques developed within the EURAMET EMRP 'MetroFission' Joint Research Project, were aimed at performing on-site activity measurements at the primary standard level (4πβ-γ coincidence counting) for a wide range of radionuclides utilizing recent advances in high-speed digital sampling and digital signal processing. The state-of-the-art technology employed within this project provides up to 14-bit digitizer systems operating with sampling rates in the order of 10 8 to 10 9 samples-per-second, incorporating on-board FPGA devices, which greatly enhances the application of digital signal processing for the implementation of digital coincidence counting. These devices when coupled to suitable analysis software, demonstrate a significant improvement in the provision of primary standards of radioactivity. This manuscript provides a description of the systems employed, along with recommendations regarding optimization of the digital sampling of signals from photo-multiplier tubes and pre-amplifiers and compare the benefits of 'off-line' versus 'on-line' 4πβ-γ digital coincidence counting systems. (authors)

  3. Method of measuring the disinteration rate of beta-emitting radionuclide in a liquid sample

    International Nuclear Information System (INIS)

    Horrocks, D.L.

    1977-01-01

    A method of measuring the disintegration rate of a beta-emitting radionuclide in a liquid sample by counting at least two differently quenched versions of the sample is described. In each counting operation the sample is counted in the presence of and in the absence of a standard radioactive source. A pulse height (PH) corresponding to a unique point on the pulse height spectrum generated in the presence of the standard is determined. A zero threshold sample count rate (CPM) is derived by counting the sample once in a counting window having a zero threshold lower limit. Normalized values of the measured pulse heights (PH) are developed and correlated with the corresponding pulse counts (CPM) to determine the pulse count for a normalized pulse height value of zero and hence the sample disintegration rate

  4. The tropical lapse rate steepened during the Last Glacial Maximum

    NARCIS (Netherlands)

    Loomis, S.E.; Russell, J.M.; Verschuren, D.; Morrill, C.; De Cort, G.; Sinninghe Damsté, J.S.; Olago, D.; Eggermont, H.; Street-Perrott, F.A.; Kelly, M.A.

    2017-01-01

    The gradient of air temperature with elevation (the temperature lapse rate) in the tropics is predicted to become lesssteep during the coming century as surface temperature rises, enhancing the threat of warming in high-mountainenvironments. However, the sensitivity of the lapse rate to climate

  5. Ceramic high-rate timing RPCs

    International Nuclear Information System (INIS)

    Lopes, L.; Ferreira Marques, R.; Fonte, P.; Hennetier, L.; Pereira, A.; Sousa Correia, A.M.

    2006-01-01

    Following some previous work, we report here considerable improvements on the counting rate capability of timing RPCs by the use of ceramic electrodes with a resistivity of 10 9 Ω.cm. The X-ray sensitivity of the detector depends linearly on the counting rate with a slope of 9% per 100 kHz/cm 2 , free from charge depletion effects, while keeping a timing accuracy, measured with 511 keV synchronous photon pairs, around 90 ps σ up to 75 kHz/cm 2

  6. Finite mixture model: A maximum likelihood estimation approach on time series data

    Science.gov (United States)

    Yen, Phoong Seuk; Ismail, Mohd Tahir; Hamzah, Firdaus Mohamad

    2014-09-01

    Recently, statistician emphasized on the fitting of finite mixture model by using maximum likelihood estimation as it provides asymptotic properties. In addition, it shows consistency properties as the sample sizes increases to infinity. This illustrated that maximum likelihood estimation is an unbiased estimator. Moreover, the estimate parameters obtained from the application of maximum likelihood estimation have smallest variance as compared to others statistical method as the sample sizes increases. Thus, maximum likelihood estimation is adopted in this paper to fit the two-component mixture model in order to explore the relationship between rubber price and exchange rate for Malaysia, Thailand, Philippines and Indonesia. Results described that there is a negative effect among rubber price and exchange rate for all selected countries.

  7. The tropical lapse rate steepened during the Last Glacial Maximum

    NARCIS (Netherlands)

    Loomis, Shannon E; Russell, James M; Verschuren, Dirk; Morrill, Carrie; De Cort, Gijs; Sinninghe Damsté, Jaap S; Olago, Daniel; Eggermont, Hilde; Street-Perrott, F Alayne; Kelly, Meredith A

    The gradient of air temperature with elevation (the temperature lapse rate) in the tropics is predicted to become less steep during the coming century as surface temperature rises, enhancing the threat of warming in high-mountain environments. However, the sensitivity of the lapse rate to climate

  8. Investigation of the Maximum Spin-Up Coefficients of Friction Obtained During Tests of a Landing Gear Having a Static-Load Rating of 20,000 Pounds

    Science.gov (United States)

    Batterson, Sidney A.

    1959-01-01

    An experimental investigation was made at the Langley landing loads track to obtain data on the maximum spin-up coefficients of friction developed by a landing gear having a static-load rating of 20,000 pounds. The forward speeds ranged from 0 to approximately 180 feet per second and the sinking speeds, from 2.7 feet per second to 9.4 feet per second. The results indicated the variation of the maximum spin-up coefficient of friction with forward speed and vertical load. Data obtained during this investigation are also compared with some results previously obtained for nonrolling tires to show the effect of forward speed.

  9. The use of counting beads to improve the classification of fast breathing in low-resource settings: a multi-country review

    Science.gov (United States)

    Noordam, Aaltje Camielle; Barberá Laínez, Yolanda; Sadruddin, Salim; van Heck, Pabla Maria; Chono, Alex Opio; Acaye, Geoffrey Larry; Lara, Victor; Nanyonjo, Agnes; Ocan, Charles; Källander, Karin

    2015-01-01

    To decrease child mortality due to common but life-threatening illnesses, community health workers (CHWs) are trained to assess, classify and treat sick children. For pneumonia, CHWs are trained to count the respiratory rate of a child with cough and/or difficulty breathing, and determine whether the child has fast breathing or not based on how the child’s breath count relates to age-specific respiratory rate cut-off points. International organizations training CHWs to classify fast breathing realized that many of them faced challenges counting and determining how the respiratory rate relates to age-specific cut-off points. Counting beads were designed to overcome these challenges. This article presents findings from different studies on the utility of these beads, in conjunction with a timer, as a tool to improve classification of fast breathing. Studies conducted by the International Rescue Committee and Save the Children among illiterate CHWs assessed the effectiveness of counting beads to improve both counting and classifying respiratory rate against age-specific cut-off points. These studies found that the use of counting beads enabled and improved the assessment and classification of fast breathing. However, a Malaria Consortium study found that the use of counting beads decreased the accuracy of counting breaths among literate CHWs. Qualitative findings from these studies and two additional studies by UNICEF suggest that the design of the beads is crucial: beads should move comfortably, and a separate bead string, with colour coding, is required for the age groups with different cut-off thresholds—eliminating more complicated calculations. Further research, using standardized protocols and gold standard comparisons, is needed to understand the accuracy of beads in comparison to other tools used for classifying pneumonia, which CHWs benefit most from each different tool (i.e. disaggregating data by levels of literacy and numeracy) and what the impact is

  10. A Burst-Mode Photon-Counting Receiver with Automatic Channel Estimation and Bit Rate Detection

    Science.gov (United States)

    2016-02-24

    Grein, M.E., Elgin, L.E., Robinson, B.S., Kachelmyer, A.L., Caplan , D.O., Stevens, M.L., Carney, J.J., Hamilton, S.A., and Boroson, D.M., “Demonstration...Robinson, B.S., Kerman, A.J., Dauler, E.A., Barron, R.J., Caplan , D.O., Stevens, M.L., Carney, J.J., Hamilton, S.A., Yang, J.K.W., and Berggren, K.K., “781...Mbit/s photon-counting optical communications using a superconducting nanowire detector,” Optics Letters, v. 31 no. 4 444-446 (2006). [14] Caplan

  11. A study of cellular counting to determine minimum thresholds for adequacy for liquid-based cervical cytology using a survey and counting protocol.

    Science.gov (United States)

    Kitchener, Henry C; Gittins, Matthew; Desai, Mina; Smith, John H F; Cook, Gary; Roberts, Chris; Turnbull, Lesley

    2015-03-01

    progressively diluted cervical samples. Laboratory practice varied in terms of threshold of cellular adequacy and of morphological markers of adequacy. While SP laboratories generally used a minimum acceptable cell count (MACC) of 15,000, the MACC employed by TP laboratories varied between 5000 and 15,000. The cell counting study showed that a standard protocol achieved moderate to strong inter-rater reproducibility. Analysis of slide reporting from laboratories revealed that a large proportion of the samples reported as inadequate had cell counts above a threshold of 15,000 for SP, and 5000 and 10,000 for TP. Inter-rater unanimity was greater among more cellular preparations. Dilution studies demonstrated greater detection of abnormalities in slides with counts above the MACC and among slides with more than 25 dyskaryotic cells. Variation in laboratory practice demonstrates a requirement for evidence-based standards for designating a MACC. This study has indicated that a MACC of 15,000 and 5000 for SP and TP, respectively, achieves a balance in terms of maintaining sensitivity and low inadequacy rates. The findings of this study should inform the development of laboratory practice guidelines. The National Institute for Health Research Health Technology Assessment programme.

  12. The influence of sovereign credit ratings on corporate credit ratings.

    OpenAIRE

    Umutoni, Liliane

    2017-01-01

    Credit ratings have a key role in modern financial markets as they communicate crucial information on the creditworthiness of a debt issuer to investors and regulators. These credit ratings are mostly determined by three rating agencies, namely Standard & Poor’s, Moody’s and Fitch, even though, the credit rating industry counts a dozen of recognized rating agencies. Indeed, the three agencies have become the market leaders with a market share of 94.3% on the U.S market (Bloomberg, 2015) and 9...

  13. Test-Retest Reliability of Rating of Perceived Exertion and Agreement With 1-Repetition Maximum in Adults.

    Science.gov (United States)

    Bove, Allyn M; Lynch, Andrew D; DePaul, Samantha M; Terhorst, Lauren; Irrgang, James J; Fitzgerald, G Kelley

    2016-09-01

    Study Design Clinical measurement. Background It has been suggested that rating of perceived exertion (RPE) may be a useful alternative to 1-repetition maximum (1RM) to determine proper resistance exercise dosage. However, the test-retest reliability of RPE for resistance exercise has not been determined. Additionally, prior research regarding the relationship between 1RM and RPE is conflicting. Objectives The purpose of this study was to (1) determine test-retest reliability of RPE related to resistance exercise and (2) assess agreement between percentages of 1RM and RPE during quadriceps resistance exercise. Methods A sample of participants with and without knee pathology completed a series of knee extension exercises and rated the perceived difficulty of each exercise on a 0-to-10 RPE scale, then repeated the procedure 1 to 2 weeks later for test-retest reliability. To determine agreement between RPE and 1RM, participants completed knee extension exercises at various percentages of their 1RM (10% to 130% of predicted 1RM) and rated the perceived difficulty of each exercise on a 0-to-10 RPE scale. Percent agreement was calculated between the 1RM and RPE at each resistance interval. Results The intraclass correlation coefficient indicated excellent test-retest reliability of RPE for quadriceps resistance exercises (intraclass correlation coefficient = 0.895; 95% confidence interval: 0.866, 0.918). Overall percent agreement between RPE and 1RM was 60%, but agreement was poor within the ranges that would typically be used for training (50% 1RM for muscle endurance, 70% 1RM and greater for strength). Conclusion Test-retest reliability of perceived exertion during quadriceps resistance exercise was excellent. However, agreement between the RPE and 1RM was poor, especially in common training zones for knee extensor strengthening. J Orthop Sports Phys Ther 2016;46(9):768-774. Epub 5 Aug 2016. doi:10.2519/jospt.2016.6498.

  14. LAWRENCE RADIATION LABORATORY COUNTING HANDBOOK

    Energy Technology Data Exchange (ETDEWEB)

    Group, Nuclear Instrumentation

    1966-10-01

    The Counting Handbook is a compilation of operational techniques and performance specifications on counting equipment in use at the Lawrence Radiation Laboratory, Berkeley. Counting notes have been written from the viewpoint of the user rather than that of the designer or maintenance man. The only maintenance instructions that have been included are those that can easily be performed by the experimenter to assure that the equipment is operating properly.

  15. Maximum parsimony, substitution model, and probability phylogenetic trees.

    Science.gov (United States)

    Weng, J F; Thomas, D A; Mareels, I

    2011-01-01

    The problem of inferring phylogenies (phylogenetic trees) is one of the main problems in computational biology. There are three main methods for inferring phylogenies-Maximum Parsimony (MP), Distance Matrix (DM) and Maximum Likelihood (ML), of which the MP method is the most well-studied and popular method. In the MP method the optimization criterion is the number of substitutions of the nucleotides computed by the differences in the investigated nucleotide sequences. However, the MP method is often criticized as it only counts the substitutions observable at the current time and all the unobservable substitutions that really occur in the evolutionary history are omitted. In order to take into account the unobservable substitutions, some substitution models have been established and they are now widely used in the DM and ML methods but these substitution models cannot be used within the classical MP method. Recently the authors proposed a probability representation model for phylogenetic trees and the reconstructed trees in this model are called probability phylogenetic trees. One of the advantages of the probability representation model is that it can include a substitution model to infer phylogenetic trees based on the MP principle. In this paper we explain how to use a substitution model in the reconstruction of probability phylogenetic trees and show the advantage of this approach with examples.

  16. To the elementary theory of critical (maximum) flow rate of two-phase mixture in channels with various sections

    International Nuclear Information System (INIS)

    Nigmatulin, B.I.; Soplenkov, K.I.

    1978-01-01

    On the basis of the concepts of two-phase dispersive flow with various structures (bubble, vapour-drop etc) in the framework of the two-speed and two-temperature one-dimension stationary model of the current with provision for phase transitions the conditions, under which a critical (maximum) flow rate of two-phase mixture is achieved during its outflowing from a channel with the pre-set geometry, have been determined. It is shown, that for the choosen set of two-phase flow equations with the known parameters of deceleration and structure one of the critical conditions is satisfied: either solution of the set of equations corresponding a critical flow rate is a special one, i.e. passes through a special point locating between minimum and outlet channel sections where the carrying phase velocity approaches the value of decelerated sound speed in the mixture or the determinator of the initial set of equations equals zero for the outlet channel sections, i.e. gradients of the main flow parameters tend to +-infinity in this section, and carrying phase velocity also approaches the value of the decelerated sound velocity in the mixture

  17. Bunch mode specific rate corrections for PILATUS3 detectors

    Energy Technology Data Exchange (ETDEWEB)

    Trueb, P., E-mail: peter.trueb@dectris.com [DECTRIS Ltd, 5400 Baden (Switzerland); Dejoie, C. [ETH Zurich, 8093 Zurich (Switzerland); Kobas, M. [DECTRIS Ltd, 5400 Baden (Switzerland); Pattison, P. [EPF Lausanne, 1015 Lausanne (Switzerland); Peake, D. J. [School of Physics, The University of Melbourne, Victoria 3010 (Australia); Radicci, V. [DECTRIS Ltd, 5400 Baden (Switzerland); Sobott, B. A. [School of Physics, The University of Melbourne, Victoria 3010 (Australia); Walko, D. A. [Argonne National Laboratory, Argonne, IL 60439 (United States); Broennimann, C. [DECTRIS Ltd, 5400 Baden (Switzerland)

    2015-04-09

    The count rate behaviour of PILATUS3 detectors has been characterized for seven bunch modes at four different synchrotrons. The instant retrigger technology of the PILATUS3 application-specific integrated circuit is found to reduce the dependency of the required rate correction on the synchrotron bunch mode. The improvement of using bunch mode specific rate corrections based on a Monte Carlo simulation is quantified. PILATUS X-ray detectors are in operation at many synchrotron beamlines around the world. This article reports on the characterization of the new PILATUS3 detector generation at high count rates. As for all counting detectors, the measured intensities have to be corrected for the dead-time of the counting mechanism at high photon fluxes. The large number of different bunch modes at these synchrotrons as well as the wide range of detector settings presents a challenge for providing accurate corrections. To avoid the intricate measurement of the count rate behaviour for every bunch mode, a Monte Carlo simulation of the counting mechanism has been implemented, which is able to predict the corrections for arbitrary bunch modes and a wide range of detector settings. This article compares the simulated results with experimental data acquired at different synchrotrons. It is found that the usage of bunch mode specific corrections based on this simulation improves the accuracy of the measured intensities by up to 40% for high photon rates and highly structured bunch modes. For less structured bunch modes, the instant retrigger technology of PILATUS3 detectors substantially reduces the dependency of the rate correction on the bunch mode. The acquired data also demonstrate that the instant retrigger technology allows for data acquisition up to 15 million photons per second per pixel.

  18. The Kruskal Count

    OpenAIRE

    Lagarias, Jeffrey C.; Rains, Eric; Vanderbei, Robert J.

    2001-01-01

    The Kruskal Count is a card trick invented by Martin J. Kruskal in which a magician "guesses" a card selected by a subject according to a certain counting procedure. With high probability the magician can correctly "guess" the card. The success of the trick is based on a mathematical principle related to coupling methods for Markov chains. This paper analyzes in detail two simplified variants of the trick and estimates the probability of success. The model predictions are compared with simula...

  19. SPERM COUNT DISTRIBUTIONS IN FERTILE MEN

    Science.gov (United States)

    Sperm concentration and count are often used as indicators of environmental impacts on male reproductive health. Existing clinical databases may be biased towards subfertile men with low sperm counts and less is known about expected sperm count distributions in cohorts of fertil...

  20. Use of Image Pro Plus for counting of α particles tracks in CR-39

    International Nuclear Information System (INIS)

    Miranda, Marcia Valeria F.E. Sa; Crispim, Verginia Reis

    2011-01-01

    This paper measured the radon radiations present in NORM samples, through the diffusion chambers which contained a nuclear track detector (CR-39). For automatic counting of those tracks the computer program Image Pro plus was used. This paper reports the application of Image Pro plus for counting the nuclear tracks, coming from the radon radiation present in NORM samples. As the radiation rate of 222 Rn if proportional to the number of these tracks, the methodology allowed to compare the levels of contamination of the analysed samples. Also, tables and graphics are presented with counting results referring to the stage of validation of Image Pro plus

  1. Method for measuring the disintegration rate of a beta-emitting radionuclide in a liquid sample

    International Nuclear Information System (INIS)

    1977-01-01

    A method of measuring the distintegration rate of a beta-emitting radionuclide in a liquid sample by counting at least two differently quenched versions of the sample. In each counting operation the sample is counted in the presence of and in the absence of a standard radioactive source. A pulse height (PH) corresponding to a unique point on the pulse height spectrum generated in the presence of the standard is determined. A zero threshold sample count rate (CPM) is derived by counting the sample once in a counting window having a zero threshold lower limit. Normalized values of the measured pulse heights (PH) are developed and correlated with the corresponding counts (CPM) to determine the pulse count for a normalized pulse height value of zero and hence the sample disintegration rate

  2. Total bacterial count and somatic cell count in refrigerated raw milk stored in communal tanks

    Directory of Open Access Journals (Sweden)

    Edmar da Costa Alves

    2014-09-01

    Full Text Available The current industry demand for dairy products with extended shelf life has resulted in new challenges for milk quality maintenance. The processing of milk with high bacterial counts compromises the quality and performance of industrial products. The study aimed to evaluate the total bacteria counts (TBC and somatic cell count (SCC in 768 samples of refrigerated raw milk, from 32 communal tanks. Samples were collected in the first quarter of 2010, 2011, 2012 and 2013 and analyzed by the Laboratory of Milk Quality - LQL. Results showed that 62.5%, 37.5%, 15.6% and 27.1% of the means for TBC in 2010, 2011, 2012 and 2013, respectively, were above the values established by legislation. However, we observed a significant reduction in the levels of total bacterial count (TBC in the studied periods. For somatic cell count, 100% of the means indicated values below 600.000 cells/mL, complying with the actual Brazilian legislation. The values found for the somatic cell count suggests the adoption of effective measures for the sanitary control of the herd. However, the results must be considered with caution as it highlights the need for quality improvements of the raw material until it achieves reliable results effectively.

  3. Usage of liquid scintillation counting for detecting the chemiluminescence of cells and its application in medicine

    International Nuclear Information System (INIS)

    Li Tianxing; Liang Qizhong; Zou Xiaowei; Yang Zhaohen; Huang Yong; Li Huaqiang

    1995-01-01

    The liquid scintillator counting-chemiluminescence (LSC-CL) of mono-photon radiance is a sensitive, handy and high-autoanalytic technique. Through measuring basic CL, dependent CL and maximum phagocytic CL of polymorphonuclear (PMN), we studied best factor levels of the method with orthogonal design [L 9 (3 4 )]. The results showed the peak forms changed markedly (inter-group P -4 M). PMN-CL in blood was measured during acute attack of the old patients with chronic bronchitis and the children with pneumonia bronchial. It was suggested that PMN phagocytosis decreased. So the dynamic analysis of maximum phagocytic CL would help us with the deep going clinical researches of the mechanisms of anti-inflammation and injuring by the oxygen free radicals

  4. Morbidity and mortality according to latest CD4+ cell count among HIV positive individuals in South Africa who enrolled in project Phidisa.

    Directory of Open Access Journals (Sweden)

    Patrick H Maduna

    Full Text Available Short-term morbidity and mortality rates for HIV positive soldiers in the South African National Defence Force (SANDF would inform decisions about deployment and HIV disease management. Risks were determined according to the latest CD4+ cell count and use of antiretroviral therapy (ART for HIV positive individuals in the SANDF and their dependents.A total of 7,114 participants were enrolled and followed for mortality over a median of 4.7 years (IQR: 1.9, 7.1 years. For a planned subset (5,976, progression of disease (POD and grade 4, potentially life-threatening events were also ascertained. CD4+ count and viral load were measured every 3 to 6 months. Poisson regression was used to compare event rates by latest CD4+ count (<50, 50-99, 100-199, 200-349, 350-499, 500+ with a focus on upper three strata, and to estimate relative risks (RRs (ART/no ART. Median entry CD4+ was 207 cells/mm3. During follow-up over 70% were prescribed ART. Over follow-up 1,226 participants died; rates ranged from 57.6 (< 50 cells to 0.8 (500+ cells per 100 person years (py. Compared to those with latest CD4+ 200-349 (2.2/100 py, death rates were significantly lower (p<0.001, as expected, for those with 350-499 (0.9/100 py and with 500+ cells (0.8/100 py. The composite outcome of death, POD or grade 4 events occurred in 2,302 participants (4,045 events; rates were similar in higher CD4+ count strata (9.4 for 350-499 and 7.9 for 500+ cells and lower than those with counts 200-349 cells (13.5 (p<0.001. For those with latest CD4+ 350+ cells, 63% of the composite outcomes (680 of 1,074 were grade 4 events.Rates of morbidity and mortality are lowest among those with CD4+ count of 350 or higher and rates do not differ for those with counts of 350-499 versus 500+ cells. Grade 4 events are the predominant morbidity for participants with CD4+ counts of 350+ cells.

  5. Analysis of photon count data from single-molecule fluorescence experiments

    Science.gov (United States)

    Burzykowski, T.; Szubiakowski, J.; Rydén, T.

    2003-03-01

    We consider single-molecule fluorescence experiments with data in the form of counts of photons registered over multiple time-intervals. Based on the observation schemes, linking back to works by Dehmelt [Bull. Am. Phys. Soc. 20 (1975) 60] and Cook and Kimble [Phys. Rev. Lett. 54 (1985) 1023], we propose an analytical approach to the data based on the theory of Markov-modulated Poisson processes (MMPP). In particular, we consider maximum-likelihood estimation. The method is illustrated using a real-life dataset. Additionally, the properties of the proposed method are investigated through simulations and compared to two other approaches developed by Yip et al. [J. Phys. Chem. A 102 (1998) 7564] and Molski [Chem. Phys. Lett. 324 (2000) 301].

  6. Rainflow counting revisited

    Energy Technology Data Exchange (ETDEWEB)

    Soeker, H [Deutsches Windenergie-Institut (Germany)

    1996-09-01

    As state of the art method the rainflow counting technique is presently applied everywhere in fatigue analysis. However, the author feels that the potential of the technique is not fully recognized in wind energy industries as it is used, most of the times, as a mere data reduction technique disregarding some of the inherent information of the rainflow counting results. The ideas described in the following aim at exploitation of this information and making it available for use in the design and verification process. (au)

  7. Radon counting statistics - a Monte Carlo investigation

    International Nuclear Information System (INIS)

    Scott, A.G.

    1996-01-01

    Radioactive decay is a Poisson process, and so the Coefficient of Variation (COV) of open-quotes nclose quotes counts of a single nuclide is usually estimated as 1/√n. This is only true if the count duration is much shorter than the half-life of the nuclide. At longer count durations, the COV is smaller than the Poisson estimate. Most radon measurement methods count the alpha decays of 222 Rn, plus the progeny 218 Po and 214 Po, and estimate the 222 Rn activity from the sum of the counts. At long count durations, the chain decay of these nuclides means that every 222 Rn decay must be followed by two other alpha decays. The total number of decays is open-quotes 3Nclose quotes, where N is the number of radon decays, and the true COV of the radon concentration estimate is 1/√(N), √3 larger than the Poisson total count estimate of 1/√3N. Most count periods are comparable to the half lives of the progeny, so the relationship between COV and count time is complex. A Monte-Carlo estimate of the ratio of true COV to Poisson estimate was carried out for a range of count periods from 1 min to 16 h and three common radon measurement methods: liquid scintillation, scintillation cell, and electrostatic precipitation of progeny. The Poisson approximation underestimates COV by less than 20% for count durations of less than 60 min

  8. Hanford whole body counting manual

    International Nuclear Information System (INIS)

    Palmer, H.E.; Rieksts, G.A.; Lynch, T.P.

    1990-06-01

    This document describes the Hanford Whole Body Counting Program as it is administered by Pacific Northwest Laboratory (PNL) in support of the US Department of Energy--Richland Operations Office (DOE-RL) and its Hanford contractors. Program services include providing in vivo measurements of internally deposited radioactivity in Hanford employees (or visitors). Specific chapters of this manual deal with the following subjects: program operational charter, authority, administration, and practices, including interpreting applicable DOE Orders, regulations, and guidance into criteria for in vivo measurement frequency, etc., for the plant-wide whole body counting services; state-of-the-art facilities and equipment used to provide the best in vivo measurement results possible for the approximately 11,000 measurements made annually; procedures for performing the various in vivo measurements at the Whole Body Counter (WBC) and related facilities including whole body counts; operation and maintenance of counting equipment, quality assurance provisions of the program, WBC data processing functions, statistical aspects of in vivo measurements, and whole body counting records and associated guidance documents. 16 refs., 48 figs., 22 tabs

  9. Image Analytical Approach for Needle-Shaped Crystal Counting and Length Estimation

    DEFF Research Database (Denmark)

    Wu, Jian X.; Kucheryavskiy, Sergey V.; Jensen, Linda G.

    2015-01-01

    Estimation of nucleation and crystal growth rates from microscopic information is of critical importance. This can be an especially challenging task if needle growth of crystals is observed. To address this challenge, an image analytical method for counting of needle-shaped crystals and estimating...

  10. Low Count Anomaly Detection at Large Standoff Distances

    Science.gov (United States)

    Pfund, David Michael; Jarman, Kenneth D.; Milbrath, Brian D.; Kiff, Scott D.; Sidor, Daniel E.

    2010-02-01

    Searching for hidden illicit sources of gamma radiation in an urban environment is difficult. Background radiation profiles are variable and cluttered with transient acquisitions from naturally occurring radioactive materials and medical isotopes. Potentially threatening sources likely will be nearly hidden in this noise and encountered at high standoff distances and low threat count rates. We discuss an anomaly detection algorithm that characterizes low count sources as threatening or non-threatening and operates well in the presence of high benign source variability. We discuss the algorithm parameters needed to reliably find sources both close to the detector and far away from it. These parameters include the cutoff frequencies of background tracking filters and the integration time of the spectrometer. This work is part of the development of the Standoff Radiation Imaging System (SORIS) as part of DNDO's Standoff Radiation Detection System Advanced Technology Demonstration (SORDS-ATD) program.

  11. Platelet Count and Plateletcrit

    African Journals Online (AJOL)

    strated that neonates with late onset sepsis (bacteremia after 3 days of age) had a dramatic increase in MPV and. PDW18. We hypothesize that as the MPV and PDW increase and platelet count and PCT decrease in sick children, intui- tively, the ratio of MPV to PCT; MPV to Platelet count,. PDW to PCT, PDW to platelet ...

  12. Neutron counting and gamma spectroscopy with PVT detectors

    International Nuclear Information System (INIS)

    Mitchell, Dean James; Brusseau, Charles A.

    2011-01-01

    Radiation portals normally incorporate a dedicated neutron counter and a gamma-ray detector with at least some spectroscopic capability. This paper describes the design and presents characterization data for a detection system called PVT-NG, which uses large polyvinyl toluene (PVT) detectors to monitor both types of radiation. The detector material is surrounded by polyvinyl chloride (PVC), which emits high-energy gamma rays following neutron capture reactions. Assessments based on high-energy gamma rays are well suited for the detection of neutron sources, particularly in border security applications, because few isotopes in the normal stream of commerce have significant gamma ray yields above 3 MeV. Therefore, an increased count rate for high-energy gamma rays is a strong indicator for the presence of a neutron source. The sensitivity of the PVT-NG sensor to bare 252 Cf is 1.9 counts per second per nanogram (cps/ng) and the sensitivity for 252 Cf surrounded by 2.5 cm of polyethylene is 2.3 cps/ng. The PVT-NG sensor is a proof-of-principal sensor that was not fully optimized. The neutron detector sensitivity could be improved, for instance, by using additional moderator. The PVT-NG detectors and associated electronics are designed to provide improved resolution, gain stability, and performance at high-count rates relative to PVT detectors in typical radiation portals. As well as addressing the needs for neutron detection, these characteristics are also desirable for analysis of the gamma-ray spectra. Accurate isotope identification results were obtained despite the common impression that the absence of photopeaks makes data collected by PVT detectors unsuitable for spectroscopic analysis. The PVT detectors in the PVT-NG unit are used for both gamma-ray and neutron detection, so the sensitive volume exceeds the volume of the detection elements in portals that use dedicated components to detect each type of radiation.

  13. Impact of marine reserve on maximum sustainable yield in a traditional prey-predator system

    Science.gov (United States)

    Paul, Prosenjit; Kar, T. K.; Ghorai, Abhijit

    2018-01-01

    Multispecies fisheries management requires managers to consider the impact of fishing activities on several species as fishing impacts both targeted and non-targeted species directly or indirectly in several ways. The intended goal of traditional fisheries management is to achieve maximum sustainable yield (MSY) from the targeted species, which on many occasions affect the targeted species as well as the entire ecosystem. Marine reserves are often acclaimed as the marine ecosystem management tool. Few attempts have been made to generalize the ecological effects of marine reserve on MSY policy. We examine here how MSY and population level in a prey-predator system are affected by the low, medium and high reserve size under different possible scenarios. Our simulation works shows that low reserve area, the value of MSY for prey exploitation is maximum when both prey and predator species have fast movement rate. For medium reserve size, our analysis revealed that the maximum value of MSY for prey exploitation is obtained when prey population has fast movement rate and predator population has slow movement rate. For high reserve area, the maximum value of MSY for prey's exploitation is very low compared to the maximum value of MSY for prey's exploitation in case of low and medium reserve. On the other hand, for low and medium reserve area, MSY for predator exploitation is maximum when both the species have fast movement rate.

  14. [Effects of HiLo for two weeks on erythrocyte immune adhesion and leukocyte count of swimmers].

    Science.gov (United States)

    Zhao, Yong-Cai; Gao, Bing-Hong; Wu, Ge-Lin; Zhang, Jiu-Li

    2012-07-01

    To investigate the effects of living high-training low (HiLo) on innate immunity in blood of elite swimmers. Six female swimmers undertook HiLo for two weeks, erythrocyte adhesion function and counts of leukocyte were tested in different time of training period. Red blood cell C3b receptor ring rate (RBC-C3bRR) decreased and red blood cell immune complex matter ring rate (RBC-ICR) increased significantly (P < 0.05), the two markers returned to base line 1 week after training. Counts of leukocyte and granulocyte decreased significantly (P < 0.05), and they recovered 1 week after training; Counts of lymphocyte and monocyte decreased without significance during training and did not recovered after training. Immunity of erythrocyte and granulocyte decreased quickly, but lymphocyte and monocyte recovered slowly, swimmers were adaptive to the training.

  15. Blood eosinophil count and exacerbations in severe chronic obstructive pulmonary disease after withdrawal of inhaled corticosteroids

    DEFF Research Database (Denmark)

    Watz, Henrik; Tetzlaff, Kay; Wouters, Emiel F M

    2016-01-01

    were seen with eosinophil cutoffs of 300 cells per μL and 400 cells per μL, and mutually exclusive subgroups. INTERPRETATION: Blood eosinophil counts at screening were related to the exacerbation rate after complete ICS withdrawal in patients with severe to very severe COPD and a history...... of exacerbations. Our data suggest that counts of 4% or greater or 300 cells per μL or more might identify a deleterious effect of ICS withdrawal, an effect not seen in most patients with eosinophil counts below these thresholds. FUNDING: Boehringer Ingelheim....

  16. Counting probe

    International Nuclear Information System (INIS)

    Matsumoto, Haruya; Kaya, Nobuyuki; Yuasa, Kazuhiro; Hayashi, Tomoaki

    1976-01-01

    Electron counting method has been devised and experimented for the purpose of measuring electron temperature and density, the most fundamental quantities to represent plasma conditions. Electron counting is a method to count the electrons in plasma directly by equipping a probe with the secondary electron multiplier. It has three advantages of adjustable sensitivity, high sensitivity of the secondary electron multiplier, and directional property. Sensitivity adjustment is performed by changing the size of collecting hole (pin hole) on the incident front of the multiplier. The probe is usable as a direct reading thermometer of electron temperature because it requires to collect very small amount of electrons, thus it doesn't disturb the surrounding plasma, and the narrow sweep width of the probe voltage is enough. Therefore it can measure anisotropy more sensitively than a Langmuir probe, and it can be used for very low density plasma. Though many problems remain on anisotropy, computer simulation has been carried out. Also it is planned to provide a Helmholtz coil in the vacuum chamber to eliminate the effect of earth magnetic field. In practical experiments, the measurement with a Langmuir probe and an emission probe mounted to the movable structure, the comparison with the results obtained in reverse magnetic field by using a Helmholtz coil, and the measurement of ionic sound wave are scheduled. (Wakatsuki, Y.)

  17. A mind you can count on: validating breath counting as a behavioral measure of mindfulness

    Directory of Open Access Journals (Sweden)

    Daniel B Levinson

    2014-10-01

    Full Text Available Mindfulness practice of present moment awareness promises many benefits, but has eluded rigorous behavioral measurement. To date, research has relied on self-reported mindfulness or heterogeneous mindfulness trainings to infer skillful mindfulness practice and its effects. In four independent studies with over 400 total participants, we present the first construct validation of a behavioral measure of mindfulness, breath counting. We found it was reliable, correlated with self-reported mindfulness, differentiated long-term meditators from age-matched controls, and was distinct from sustained attention and working memory measures. In addition, we employed breath counting to test the nomological network of mindfulness. As theorized, we found skill in breath counting associated with more meta-awareness, less mind wandering, better mood, and greater nonattachment (i.e. less attentional capture by distractors formerly paired with reward. We also found in a randomized online training study that 4 weeks of breath counting training improved mindfulness and decreased mind wandering relative to working memory training and no training controls. Together, these findings provide the first evidence for breath counting as a behavioral measure of mindfulness.

  18. The maximum entropy production and maximum Shannon information entropy in enzyme kinetics

    Science.gov (United States)

    Dobovišek, Andrej; Markovič, Rene; Brumen, Milan; Fajmut, Aleš

    2018-04-01

    We demonstrate that the maximum entropy production principle (MEPP) serves as a physical selection principle for the description of the most probable non-equilibrium steady states in simple enzymatic reactions. A theoretical approach is developed, which enables maximization of the density of entropy production with respect to the enzyme rate constants for the enzyme reaction in a steady state. Mass and Gibbs free energy conservations are considered as optimization constraints. In such a way computed optimal enzyme rate constants in a steady state yield also the most uniform probability distribution of the enzyme states. This accounts for the maximal Shannon information entropy. By means of the stability analysis it is also demonstrated that maximal density of entropy production in that enzyme reaction requires flexible enzyme structure, which enables rapid transitions between different enzyme states. These results are supported by an example, in which density of entropy production and Shannon information entropy are numerically maximized for the enzyme Glucose Isomerase.

  19. Particle Count Limits Recommendation for Aviation Fuel

    Science.gov (United States)

    2015-10-05

    Particle Counter Methodology • Particle counts are taken utilizing calibration methodologies and standardized cleanliness code ratings – ISO 11171 – ISO...Limits Receipt Vehicle Fuel Tank Fuel Injector Aviation Fuel DEF (AUST) 5695B 18/16/13 Parker 18/16/13 14/10/7 Pamas / Parker / Particle Solutions 19/17...12 U.S. DOD 19/17/14/13* Diesel Fuel World Wide Fuel Charter 5th 18/16/13 DEF (AUST) 5695B 18/16/13 Caterpillar 18/16/13 Detroit Diesel 18/16/13 MTU

  20. Stability and reproducibility of gel-suspension samples for the liquid scintillation counting of 14C using N-lauroyl-L-glutamic-α,γ-dibutylamide

    International Nuclear Information System (INIS)

    Wakabayashi, G.; Ohura, H.; Okai, T.; Matoba, M.

    1999-01-01

    Stability and reproducibility of gel-suspension method for 14 C activity measurement. Commercially available gelling agent, N-lauroyl-L-glutamic-α,γ-dibutylamide, was used for the gel-formation of the samples. No change of the counting rate for the gel-suspension sample was observed for more than 2 years after the sample preparation. Four samples used for checking the reproducibility of the sample preparation method. The same values were obtained for the counting rate of 14 C activity within the counting error. No change of the counting rate was observed for the 're-gelated' sample. These results show that the gel-suspension method is appropriate for the 14 C activity measurement by the liquid scintillation method and is useful for a long-term preservation of the sample for repeated measurement. (author)

  1. Maximum-Entropy Inference with a Programmable Annealer

    Science.gov (United States)

    Chancellor, Nicholas; Szoke, Szilard; Vinci, Walter; Aeppli, Gabriel; Warburton, Paul A.

    2016-03-01

    Optimisation problems typically involve finding the ground state (i.e. the minimum energy configuration) of a cost function with respect to many variables. If the variables are corrupted by noise then this maximises the likelihood that the solution is correct. The maximum entropy solution on the other hand takes the form of a Boltzmann distribution over the ground and excited states of the cost function to correct for noise. Here we use a programmable annealer for the information decoding problem which we simulate as a random Ising model in a field. We show experimentally that finite temperature maximum entropy decoding can give slightly better bit-error-rates than the maximum likelihood approach, confirming that useful information can be extracted from the excited states of the annealer. Furthermore we introduce a bit-by-bit analytical method which is agnostic to the specific application and use it to show that the annealer samples from a highly Boltzmann-like distribution. Machines of this kind are therefore candidates for use in a variety of machine learning applications which exploit maximum entropy inference, including language processing and image recognition.

  2. 106Ru and 125I radiation dose rate gauge

    International Nuclear Information System (INIS)

    Machaj, B.; Swistowski, E.; Do Hoang Cuong

    2002-01-01

    Pulse count rate from plastic scintillator is a measure of the dose rate. Low dead time of measured channel and digital processing of measuring head signal with compensation of dead time enables correct registration of very high pulse count rate. The radiation source is set with an accuracy not worse than 0.1 mm in relation to the scintillator, and the movement of the source in horizontal and vertical direction is done with the accuracy of 0.01 mm. Additionally the gauge permits to measure the source activity and to check the uniform distribution of the radioactive material on the source surface. Random error due to pulse count rate fluctuation is negligible. The error due to instability of PTM gain is approx. 1,5% for 106 Ru and 5% for 125 I. (author)

  3. Total number of tillers of different accessions of Panicum maximum Jacq.

    Directory of Open Access Journals (Sweden)

    Thiago Perez Granato

    2012-12-01

    Full Text Available The productivity of forage grasses is due to continuous emission of leaves and tillers, ensuring the restoration of leaf área after cutting or grazing, thus ensuring the sustainability of forage. This study aimed to asses the total number tillers in different acessions of Panicum maximum Jacq. The experiment was carried in field belonging to the Instituto de Zootecnia located in Nova Odessa / SP. Evaluated two new accesses Panicum maximum, and two commercial cultivars. The cultivars tested were Aruana, Milenio, NO 2487, NO 78, and the two latter belonging to the Germoplasm Collection of the IZ. The experimental desing was in randomized complete block with four replications. The experimental area consisted of 16 plots of 10 m2 (5 x 2 m each. The experimental area was analyzed and according to the results, received dolomitic limestone corresponding 2t /ha, two months before the implementation of the experiment. Sowing was made by broad costing together with 80 kg/ha of P2O5 in the form of single superphosfate. After 60 days of implantation of the experiment it was a made a leveling of the plots to a height of about 15 cm. After this it was applied 250g of the 20-00-20 fertilizer/plot. Thirty days after the standardization it was evaluated the total number of tillers of the cultivars, using a metal frame of 0.5 x 0.5m which was thrown at random on each of the 16 plots, leaving one meter of each extremitly, and all tillers which were within the frame counted. After finished the counting of all tillers, the plots cut again at a height of approximately 15 cm. The second evaluation took place after thirty days, and it was again counted the total number of tillers following the same procedure. The results were analyzed by Tukey test at 5% after transforming the data to log(x. For the first evaluation there was no statistical difference in the total number of tillers between cultivars. But, in the second evaluation, the total number of tillers of NO 78

  4. Comparison of pregnancy rates in pre-treatment male infertility and low total motile sperm count at insemination.

    Science.gov (United States)

    Xiao, Cheng Wei; Agbo, Chioma; Dahan, Michael H

    2016-01-01

    In intrauterine insemination (IUI), total motile sperm count (TMSC) is an important predictor of pregnancy. However, the clinical significance of a poor TMSC on the day of IUI in a patient with prior normal semen analysis (SA) is unclear. We performed this study to determine if these patients perform as poorly as those who had male factor infertility diagnosed prior to commencing treatment. 147 males with two abnormal SA based on the 2010 World Health Organization criteria underwent 356 IUI with controlled ovarian hyper-stimulation (COH). Their pregnancy rates were compared to 120 males who had abnormal TMSC at the time of 265 IUI with COH, in a retrospective university-based study. The two groups were comparable in female age (p = 0.11), duration of infertility (p = 0.17), previous pregnancies (p = 0.13), female basal serum FSH level (p = 0.54) and number of mature follicles on the day of ovulation trigger (p = 0.27). Despite better semen parameters on the day of IUI in the pre-treatment male factor infertility group (TMSC mean ± SD: 61 ± 30 million vs. 3.5 ± 2 million, p male factor infertility. More studies should be performed to confirm these findings.

  5. Photon-counting 1.0 GHz-phase-modulation fluorometer

    International Nuclear Information System (INIS)

    Mizuno, T.; Nakao, S.; Mizutani, Y.; Iwata, T.

    2015-01-01

    We have constructed an improved version of a photon-counting phase-modulation fluorometer (PC-PMF) with a maximum modulation frequency of 1.0 GHz, where a phase domain measurement is conducted with a time-correlated single-photon-counting electronics. While the basic concept of the PC-PMF has been reported previously by one of the authors, little attention has been paid to its significance, other than its weak fluorescence measurement capability. Recently, we have recognized the importance of the PC-PMF and its potential for fluorescence lifetime measurements. One important aspect of the PC-PMF is that it enables us to perform high-speed measurements that exceed the frequency bandwidths of the photomultiplier tubes that are commonly used as fluorescence detectors. We describe the advantages of the PC-PMF and demonstrate its usefulness based on fundamental performance tests. In our new version of the PC-PMF, we have used a laser diode (LD) as an excitation light source rather than the light-emitting diode that was used in the primary version. We have also designed a simple and stable LD driver to modulate the device. Additionally, we have obtained a sinusoidal histogram waveform that has multiple cycles within a time span to be measured, which is indispensable for precise phase measurements. With focus on the fluorescence intensity and the resolution time, we have compared the performance of the PC-PMF with that of a conventional PMF using the analogue light detection method

  6. Photon-counting 1.0 GHz-phase-modulation fluorometer

    Energy Technology Data Exchange (ETDEWEB)

    Mizuno, T.; Nakao, S.; Mizutani, Y.; Iwata, T., E-mail: iwata@tokushima-u.ac.jp [Division of Energy System, Institute of Technology and Science, Tokushima University, 2-1 Minami-Jyosanjima, Tokushima 770-8506 (Japan)

    2015-04-15

    We have constructed an improved version of a photon-counting phase-modulation fluorometer (PC-PMF) with a maximum modulation frequency of 1.0 GHz, where a phase domain measurement is conducted with a time-correlated single-photon-counting electronics. While the basic concept of the PC-PMF has been reported previously by one of the authors, little attention has been paid to its significance, other than its weak fluorescence measurement capability. Recently, we have recognized the importance of the PC-PMF and its potential for fluorescence lifetime measurements. One important aspect of the PC-PMF is that it enables us to perform high-speed measurements that exceed the frequency bandwidths of the photomultiplier tubes that are commonly used as fluorescence detectors. We describe the advantages of the PC-PMF and demonstrate its usefulness based on fundamental performance tests. In our new version of the PC-PMF, we have used a laser diode (LD) as an excitation light source rather than the light-emitting diode that was used in the primary version. We have also designed a simple and stable LD driver to modulate the device. Additionally, we have obtained a sinusoidal histogram waveform that has multiple cycles within a time span to be measured, which is indispensable for precise phase measurements. With focus on the fluorescence intensity and the resolution time, we have compared the performance of the PC-PMF with that of a conventional PMF using the analogue light detection method.

  7. DC KIDS COUNT e-Databook Indicators

    Science.gov (United States)

    DC Action for Children, 2012

    2012-01-01

    This report presents indicators that are included in DC Action for Children's 2012 KIDS COUNT e-databook, their definitions and sources and the rationale for their selection. The indicators for DC KIDS COUNT represent a mix of traditional KIDS COUNT indicators of child well-being, such as the number of children living in poverty, and indicators of…

  8. Maximum Smoke Temperature in Non-Smoke Model Evacuation Region for Semi-Transverse Tunnel Fire

    OpenAIRE

    B. Lou; Y. Qiu; X. Long

    2017-01-01

    Smoke temperature distribution in non-smoke evacuation under different mechanical smoke exhaust rates of semi-transverse tunnel fire were studied by FDS numerical simulation in this paper. The effect of fire heat release rate (10MW 20MW and 30MW) and exhaust rate (from 0 to 160m3/s) on the maximum smoke temperature in non-smoke evacuation region was discussed. Results show that the maximum smoke temperature in non-smoke evacuation region decreased with smoke exhaust rate. Plug-holing was obse...

  9. In vivo counting of uranium

    International Nuclear Information System (INIS)

    Palmer, H.E.

    1985-03-01

    A state-of-the-art radiation detector system consisting of six individually mounted intrinsic germanium planar detectors, each 20 cm 2 by 13 mm thick, mounted together such that the angle of the whole system can be changed to match the slope of the chest of the person being counted, is described. The sensitivity of the system for counting uranium and plutonium in vivo and the precedures used in calibrating the system are also described. Some results of counts done on uranium mill workers are presented. 15 figs., 2 tabs

  10. Evaluation of left ventricular function in patients with atrial fibrillation by ECG gated blood pool scintigraphy. Using frame count normalization method

    Energy Technology Data Exchange (ETDEWEB)

    Akanabe, Hiroshi; Oshima, Motoo; Sakuma, Sadayuki

    1988-07-01

    The assumption necessary to perform ECG gated blood pool scintigraphy (EGBPS) are seemingly not valid for patients with atrial fibrillation (af), since they have wide variability in cardiac cycle length. The data were acquired in frame mode within the limits of mean heart rate of fix the first diastolic volume, and were calculated by frame count normalization (FCN) method to correct total counts in each frame. EGBPS were performed twelve patients with af, who were operated against valvular disease. The data acquired within mean heart rate +-10 % in frame mode were divided to 32 frames, and calculated total frame counts. With FCN method total frame counts from at 22nd to 32nd frame were multiplied to be equal to the average of total frame counts. FCN method could correct total frame counts at the latter frames. And there was good correlation between left ventricular ejection fraction calculated from scintigraphy and that from contrast cineangiography. Thus EGBPS with FCN method may be allow estimation of cardiac function even in subjects with af.

  11. Alpha scintillation radon counting

    International Nuclear Information System (INIS)

    Lucas, H.F. Jr.

    1977-01-01

    Radon counting chambers which utilize the alpha-scintillation properties of silver activated zinc sulfide are simple to construct, have a high efficiency, and, with proper design, may be relatively insensitive to variations in the pressure or purity of the counter filling. Chambers which were constructed from glass, metal, or plastic in a wide variety of shapes and sizes were evaluated for the accuracy and the precision of the radon counting. The principles affecting the alpha-scintillation radon counting chamber design and an analytic system suitable for a large scale study of the 222 Rn and 226 Ra content of either air or other environmental samples are described. Particular note is taken of those factors which affect the accuracy and the precision of the method for monitoring radioactivity around uranium mines

  12. Maximum Aerobic Capacity of Underground Coal Miners in India

    Directory of Open Access Journals (Sweden)

    Ratnadeep Saha

    2011-01-01

    Full Text Available Miners fitness test was assessed in terms of determination of maximum aerobic capacity by an indirect method following a standard step test protocol before going down to mine by taking into consideration of heart rates (Telemetric recording and oxygen consumption of the subjects (Oxylog-II during exercise at different working rates. Maximal heart rate was derived as 220−age. Coal miners reported a maximum aerobic capacity within a range of 35–38.3 mL/kg/min. It also revealed that oldest miners (50–59 yrs had a lowest maximal oxygen uptake (34.2±3.38 mL/kg/min compared to (42.4±2.03 mL/kg/min compared to (42.4±2.03 mL/kg/min the youngest group (20–29 yrs. It was found to be negatively correlated with age (r=−0.55 and −0.33 for younger and older groups respectively and directly associated with the body weight of the subjects (r=0.57 – 0.68, P≤0.001. Carriers showed maximum cardio respiratory capacity compared to other miners. Indian miners VO2max was found to be lower both compared to their abroad mining counterparts and various other non-mining occupational working groups in India.

  13. High rate spectroscopy for on-line nuclear coal analyzer (Nucoalyzer)

    International Nuclear Information System (INIS)

    McQuaid, J.H.; Brown, D.R.; Gozani, T.; Bozorgmanesh, H.

    1980-01-01

    A high count rate, time-variant Ge(Li) spectrometer has been developed for on-line coal analysis. The analyzer is being fabricated for use in a power generating station. Prompt neutron activation of coal samples is the basis of analysis, with 252 Cf as the source for irradiation. The spectroscopy system allows counting rates up to 150 k counts per second without significant loss in energy resolution or peak shape. The high data throughput allows the coal analyzer to be used for on-line process control. The coal analyzer will be discussed, with emphasis on the high-rate signal processing system. Results of analysis of coal samples will be presented

  14. Effect of dopamine injection on the hemocyte count and prophenoloxidase system of the white shrimp Litopenaeus vannamei

    Science.gov (United States)

    Pan, Luqing; Hu, Fawen; Zheng, Debin

    2011-09-01

    Effects of dopamine injection on the hemocyte count, phenoloxidase activity, serine proteinase activity, proteinase inhibitor activity and α2-macroglobulin-like activity in L. vannamei were studied. Results showed that dopamine injection resulted in a significant effect on the parameters measured ( P < 0.05), while no significant difference was observed in the control group (0.85% NaCl). In the experimental groups, the hemocyte count reached the minimum in 3 h; granular and semi-granular cells became stable after 12 h and hyaline cells and the total hemocyte count became stable after 18 h. Phenoloxidase activity reached the minimum in 6 h, and then became stable after 9 h. Serine protease activity and proteinase inhibitor activity reached the minimum in 3 h, and α2-macroglobulin-like activity reached the maximum in 3 h, and all the three parameters became stable after 12 h. The results suggest that the activating mechanisms of the proPO system triggered by dopamine are different from those triggered by invasive agents or spontaneously activated under a normal physical condition.

  15. Blood Count Tests: MedlinePlus Health Topic

    Science.gov (United States)

    ... Spanish WBC count (Medical Encyclopedia) Also in Spanish Topic Image MedlinePlus Email Updates Get Blood Count Tests ... WBC count Show More Show Less Related Health Topics Bleeding Disorders Blood Laboratory Tests National Institutes of ...

  16. Evaluation of lactate, white blood cell count, neutrophil count, procalcitonin and immature granulocyte count as biomarkers for sepsis in emergency department patients.

    Science.gov (United States)

    Karon, Brad S; Tolan, Nicole V; Wockenfus, Amy M; Block, Darci R; Baumann, Nikola A; Bryant, Sandra C; Clements, Casey M

    2017-11-01

    Lactate, white blood cell (WBC) and neutrophil count, procalcitonin and immature granulocyte (IG) count were compared for the prediction of sepsis, and severe sepsis or septic shock, in patients presenting to the emergency department (ED). We prospectively enrolled 501 ED patients with a sepsis panel ordered for suspicion of sepsis. WBC, neutrophil, and IG counts were measured on a Sysmex XT-2000i analyzer. Lactate was measured by i-STAT, and procalcitonin by Brahms Kryptor. We classified patients as having sepsis using a simplification of the 1992 consensus conference sepsis definitions. Patients with sepsis were further classified as having severe sepsis or septic shock using established criteria. Univariate receiver operating characteristic (ROC) analysis was performed to determine odds ratio (OR), area under the ROC curve (AUC), and sensitivity/specificity at optimal cut-off for prediction of sepsis (vs. no sepsis), and prediction of severe sepsis or septic shock (vs. no sepsis). There were 267 patients without sepsis; and 234 with sepsis, including 35 patients with severe sepsis or septic shock. Lactate had the highest OR (1.44, 95th% CI 1.20-1.73) for the prediction of sepsis; while WBC, neutrophil count and percent (neutrophil/WBC) had OR>1.00 (psepsis or septic shock, with an odds ratio (95th% CI) of 2.70 (2.02-3.61) and AUC 0.89 (0.82-0.96). Traditional biomarkers (lactate, WBC, neutrophil count, procalcitonin, IG) have limited utility in the prediction of sepsis. Copyright © 2017 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

  17. Counting and confusion: Bayesian rate estimation with multiple populations

    Science.gov (United States)

    Farr, Will M.; Gair, Jonathan R.; Mandel, Ilya; Cutler, Curt

    2015-01-01

    We show how to obtain a Bayesian estimate of the rates or numbers of signal and background events from a set of events when the shapes of the signal and background distributions are known, can be estimated, or approximated; our method works well even if the foreground and background event distributions overlap significantly and the nature of any individual event cannot be determined with any certainty. We give examples of determining the rates of gravitational-wave events in the presence of background triggers from a template bank when noise parameters are known and/or can be fit from the trigger data. We also give an example of determining globular-cluster shape, location, and density from an observation of a stellar field that contains a nonuniform background density of stars superimposed on the cluster stars.

  18. Serum Copper Level Significantly Influences Platelet Count, Lymphocyte Count and Mean Cell Hemoglobin in Sickle Cell Anemia

    Directory of Open Access Journals (Sweden)

    Okocha Chide

    2015-12-01

    Full Text Available Background Changes in serum micro nutrients levels affect a number of critically important metabolic processes; these could potentially influence blood counts and ultimately disease presentation in patients with sickle cell anemia (SCA. Objectives To evaluate the influence of serum micro-nutrients levels; zinc, copper, selenium and magnesium on blood counts in steady state SCA patients. Methods A cross sectional study that involved 28 steady state adult SCA subjects. Seven milliliters (mls of blood was collected; 3 mls was for hemoglobin electrophoresis and full blood count determination while 4 mls was for measurement of serum micro nutrients levels, by the atomic absorption spectrophotometry. Correlation between serum micro-nutrient levels and blood counts was done by the Pearson’s linear regression. Ethical approval was obtained from the institutional review board and each participant gave informed consent. All data was analyzed by SPSS software version 20. Results There was a significant correlation between serum copper levels and mean cell hemoglobin (MCH, platelet and lymphocyte counts (r = 0.418; P = 0.02, r = -0.376; P = 0.04 and r = -0.383; P = 0.04, respectively. There were no significant correlations between serum levels of other micro nutrients (selenium, zinc and magnesium and blood counts. Conclusions Copper influences blood count in SCA patients probably by inducing red cell haemolysis, oxidant tissue damage and stimulating the immune system.

  19. Method for measuring the decay rate of a radionuclide emitting β-rays in a liquid sample

    International Nuclear Information System (INIS)

    Horrocks, D.

    1977-01-01

    With this method the decay rate of a radionuclide emitting β-rays, e.g. 3 H or 14 C, in a liquid sample can be measured by means of liquid scintillation counters, at least two different versions of the sample being used with quench effect (shifting of the Compton spectrum). For this purpose each sample is counted with and without a radioactive standard source, e.g. 137 Cs. Then a pulse height will be determined corresponding to a selected point in the pulse height spectrum if the standard source is present. The determination of a zero-threshold sample count rate is then performed by counting the sample in a counting window. In addition standardized values of the measured pulse heights are derived and put in mathematical relation to corresponding pulse count rates, the pulse count rate for a standardized pulse height value thus becoming zero and the sample decay rate in this way being determined. (DG) 891 HP [de

  20. Database crime to crime match rate calculation.

    Science.gov (United States)

    Buckleton, John; Bright, Jo-Anne; Walsh, Simon J

    2009-06-01

    Guidance exists on how to count matches between samples in a crime sample database but we are unable to locate a definition of how to estimate a match rate. We propose a method that does not proceed from the match counting definition but which has a strong logic.

  1. Approximate maximum parsimony and ancestral maximum likelihood.

    Science.gov (United States)

    Alon, Noga; Chor, Benny; Pardi, Fabio; Rapoport, Anat

    2010-01-01

    We explore the maximum parsimony (MP) and ancestral maximum likelihood (AML) criteria in phylogenetic tree reconstruction. Both problems are NP-hard, so we seek approximate solutions. We formulate the two problems as Steiner tree problems under appropriate distances. The gist of our approach is the succinct characterization of Steiner trees for a small number of leaves for the two distances. This enables the use of known Steiner tree approximation algorithms. The approach leads to a 16/9 approximation ratio for AML and asymptotically to a 1.55 approximation ratio for MP.

  2. ON THE MAXIMUM MASS OF STELLAR BLACK HOLES

    International Nuclear Information System (INIS)

    Belczynski, Krzysztof; Fryer, Chris L.; Bulik, Tomasz; Ruiter, Ashley; Valsecchi, Francesca; Vink, Jorick S.; Hurley, Jarrod R.

    2010-01-01

    We present the spectrum of compact object masses: neutron stars and black holes (BHs) that originate from single stars in different environments. In particular, we calculate the dependence of maximum BH mass on metallicity and on some specific wind mass loss rates (e.g., Hurley et al. and Vink et al.). Our calculations show that the highest mass BHs observed in the Galaxy M bh ∼ 15 M sun in the high metallicity environment (Z = Z sun = 0.02) can be explained with stellar models and the wind mass loss rates adopted here. To reach this result we had to set luminous blue variable mass loss rates at the level of ∼10 -4 M sun yr -1 and to employ metallicity-dependent Wolf-Rayet winds. With such winds, calibrated on Galactic BH mass measurements, the maximum BH mass obtained for moderate metallicity (Z = 0.3 Z sun = 0.006) is M bh,max = 30 M sun . This is a rather striking finding as the mass of the most massive known stellar BH is M bh = 23-34 M sun and, in fact, it is located in a small star-forming galaxy with moderate metallicity. We find that in the very low (globular cluster-like) metallicity environment the maximum BH mass can be as high as M bh,max = 80 M sun (Z = 0.01 Z sun = 0.0002). It is interesting to note that X-ray luminosity from Eddington-limited accretion onto an 80 M sun BH is of the order of ∼10 40 erg s -1 and is comparable to luminosities of some known ultra-luminous X-ray sources. We emphasize that our results were obtained for single stars only and that binary interactions may alter these maximum BH masses (e.g., accretion from a close companion). This is strictly a proof-of-principle study which demonstrates that stellar models can naturally explain even the most massive known stellar BHs.

  3. Interpretation of galaxy counts

    International Nuclear Information System (INIS)

    Tinsely, B.M.

    1980-01-01

    New models are presented for the interpretation of recent counts of galaxies to 24th magnitude, and predictions are shown to 28th magnitude for future comparison with data from the Space Telescope. The results supersede earlier, more schematic models by the author. Tyson and Jarvis found in their counts a ''local'' density enhancement at 17th magnitude, on comparison with the earlier models; the excess is no longer significant when a more realistic mixture of galaxy colors is used. Bruzual and Kron's conclusion that Kron's counts show evidence for evolution at faint magnitudes is confirmed, and it is predicted that some 23d magnitude galaxies have redshifts greater than unity. These may include spheroidal systems, elliptical galaxies, and the bulges of early-type spirals and S0's, seen during their primeval rapid star formation

  4. Extension of the Dytlewski-style dead time correction formalism for neutron multiplicity counting to any order

    International Nuclear Information System (INIS)

    Croft, Stephen; Favalli, Andrea

    2017-01-01

    Here, neutron multiplicity counting using shift-register calculus is an established technique in the science of international nuclear safeguards for the identification, verification, and assay of special nuclear materials. Typically passive counting is used for Pu and mixed Pu-U items and active methods are used for U materials. Three measured counting rates, singles, doubles and triples are measured and, in combination with a simple analytical point-model, are used to calculate characteristics of the measurement item in terms of known detector and nuclear parameters. However, the measurement problem usually involves more than three quantities of interest, but even in cases where the next higher order count rate, quads, is statistically viable, it is not quantitatively applied because corrections for dead time losses are currently not available in the predominant analysis paradigm. In this work we overcome this limitation by extending the commonly used dead time correction method, developed by Dytlewski, to quads. We also give results for pents, which may be of interest for certain special investigations. Extension to still higher orders, may be accomplished by inspection based on the sequence presented. We discuss the foundations of the Dytlewski method, give limiting cases, and highlight the opportunities and implications that these new results expose. In particular there exist a number of ways in which the new results may be combined with other approaches to extract the correlated rates, and this leads to various practical implementations.

  5. n-Order and maximum fuzzy similarity entropy for discrimination of signals of different complexity: Application to fetal heart rate signals.

    Science.gov (United States)

    Zaylaa, Amira; Oudjemia, Souad; Charara, Jamal; Girault, Jean-Marc

    2015-09-01

    This paper presents two new concepts for discrimination of signals of different complexity. The first focused initially on solving the problem of setting entropy descriptors by varying the pattern size instead of the tolerance. This led to the search for the optimal pattern size that maximized the similarity entropy. The second paradigm was based on the n-order similarity entropy that encompasses the 1-order similarity entropy. To improve the statistical stability, n-order fuzzy similarity entropy was proposed. Fractional Brownian motion was simulated to validate the different methods proposed, and fetal heart rate signals were used to discriminate normal from abnormal fetuses. In all cases, it was found that it was possible to discriminate time series of different complexity such as fractional Brownian motion and fetal heart rate signals. The best levels of performance in terms of sensitivity (90%) and specificity (90%) were obtained with the n-order fuzzy similarity entropy. However, it was shown that the optimal pattern size and the maximum similarity measurement were related to intrinsic features of the time series. Copyright © 2015 Elsevier Ltd. All rights reserved.

  6. Comparing avian and bat fatality rate estimates among North American wind energy projects

    Energy Technology Data Exchange (ETDEWEB)

    Smallwood, Shawn

    2011-07-01

    Full text: Wind energy development has expanded rapidly, and so have concerns over bird and bat impacts caused by wind turbines. To assess and compare impacts due to collisions, investigators use a common metric, fatalities/MW/year, but estimates of fatality rates have come from various wind turbine models, tower heights, environments, fatality search methods, and analytical methods. To improve comparability and asses large-scale impacts, I applied a common set of assumptions and methods to data in fatality monitoring reports to estimate fatality rates of birds and bats at 71 wind projects across North America (52 outside the Altamont Pass Wind Resource Area, APWRA). The data were from wind turbines of 27 sizes (range 0.04-3.00 MW) and 28 tower heights (range 18.5-90 m), and searched at 40 periodic intervals (range 1-90 days) and out to 20 distances from turbines (range 30-126 m). Estimates spanned the years 1982 to 2010, and involved 1-1,345 turbines per unique combination of project, turbine size, tower height, and search methodology. I adjusted fatality rates for search detection rates averaged from 425 detection trials, and for scavenger removal rates based on 413 removal trials. I also adjusted fatality rates for turbine tower height and maximum search radius, based on logistic functions fit to cumulative counts of carcasses that were detected at 1-m distance intervals from the turbine. For each tower height, I estimated the distance at which cumulative carcass counts reached an asymptote, and for each project I calculated the proportion of fatalities likely not found due to the maximum search radius being short of the model-predicted distance asymptote. I used the same estimator in all cases. I estimated mean fatalities/MW/year among North American wind projects at 12.6 bats (80% CI: 8.1-17.1) and 11.1 birds (80% CI: 9.5-12.7), including 1.6 raptors (80% CI: 1.3-2.0), and excluding the Altamont Pass I estimated fatality rates at 17.2 bats (80% CI: 9

  7. A comparison of methods of predicting maximum oxygen uptake.

    OpenAIRE

    Grant, S; Corbett, K; Amjad, A M; Wilson, J; Aitchison, T

    1995-01-01

    The aim of this study was to compare the results from a Cooper walk run test, a multistage shuttle run test, and a submaximal cycle test with the direct measurement of maximum oxygen uptake on a treadmill. Three predictive tests of maximum oxygen uptake--linear extrapolation of heart rate of VO2 collected from a submaximal cycle ergometer test (predicted L/E), the Cooper 12 min walk, run test, and a multi-stage progressive shuttle run test (MST)--were performed by 22 young healthy males (mean...

  8. A Frank mixture copula family for modeling higher-order correlations of neural spike counts

    International Nuclear Information System (INIS)

    Onken, Arno; Obermayer, Klaus

    2009-01-01

    In order to evaluate the importance of higher-order correlations in neural spike count codes, flexible statistical models of dependent multivariate spike counts are required. Copula families, parametric multivariate distributions that represent dependencies, can be applied to construct such models. We introduce the Frank mixture family as a new copula family that has separate parameters for all pairwise and higher-order correlations. In contrast to the Farlie-Gumbel-Morgenstern copula family that shares this property, the Frank mixture copula can model strong correlations. We apply spike count models based on the Frank mixture copula to data generated by a network of leaky integrate-and-fire neurons and compare the goodness of fit to distributions based on the Farlie-Gumbel-Morgenstern family. Finally, we evaluate the importance of using proper single neuron spike count distributions on the Shannon information. We find notable deviations in the entropy that increase with decreasing firing rates. Moreover, we find that the Frank mixture family increases the log likelihood of the fit significantly compared to the Farlie-Gumbel-Morgenstern family. This shows that the Frank mixture copula is a useful tool to assess the importance of higher-order correlations in spike count codes.

  9. Maximum Urine Flow Rate of Less than 15ml/Sec Increasing Risk of Urine Retention and Prostate Surgery among Patients with Alpha-1 Blockers: A 10-Year Follow Up Study.

    Directory of Open Access Journals (Sweden)

    Hsin-Ho Liu

    Full Text Available The aim of this study was to determine the subsequent risk of acute urine retention and prostate surgery in patients receiving alpha-1 blockers treatment and having a maximum urinary flow rate of less than 15ml/sec.We identified patients who were diagnosed with benign prostate hyperplasia (BPH and had a maximum uroflow rate of less than 15ml/sec between 1 January, 2002 to 31 December, 2011 from Taiwan's National Health Insurance Research Database into study group (n = 303. The control cohort included four BPH/LUTS patients without 5ARI used for each study group, randomly selected from the same dataset (n = 1,212. Each patient was monitored to identify those who subsequently developed prostate surgery and acute urine retention.Prostate surgery and acute urine retention are detected in 5.9% of control group and 8.3% of study group during 10-year follow up. Compared with the control group, there was increase in the risk of prostate surgery and acute urine retention in the study group (HR = 1.83, 95% CI: 1.16 to 2.91 after adjusting for age, comorbidities, geographic region and socioeconomic status.Maximum urine flow rate of less than 15ml/sec is a risk factor of urinary retention and subsequent prostate surgery in BPH patients receiving alpha-1 blocker therapy. This result can provide a reference for clinicians.

  10. Recursive algorithms for phylogenetic tree counting.

    Science.gov (United States)

    Gavryushkina, Alexandra; Welch, David; Drummond, Alexei J

    2013-10-28

    In Bayesian phylogenetic inference we are interested in distributions over a space of trees. The number of trees in a tree space is an important characteristic of the space and is useful for specifying prior distributions. When all samples come from the same time point and no prior information available on divergence times, the tree counting problem is easy. However, when fossil evidence is used in the inference to constrain the tree or data are sampled serially, new tree spaces arise and counting the number of trees is more difficult. We describe an algorithm that is polynomial in the number of sampled individuals for counting of resolutions of a constraint tree assuming that the number of constraints is fixed. We generalise this algorithm to counting resolutions of a fully ranked constraint tree. We describe a quadratic algorithm for counting the number of possible fully ranked trees on n sampled individuals. We introduce a new type of tree, called a fully ranked tree with sampled ancestors, and describe a cubic time algorithm for counting the number of such trees on n sampled individuals. These algorithms should be employed for Bayesian Markov chain Monte Carlo inference when fossil data are included or data are serially sampled.

  11. ORIGINAL ARTICLES Surgical practice in a maximum security prison

    African Journals Online (AJOL)

    Prison Clinic, Mangaung Maximum Security Prison, Bloemfontein. F Kleinhans, BA (Cur) .... HIV positivity rate and the use of the rectum to store foreign objects. ... fruit in sunlight. Other positive health-promoting factors may also play a role,.

  12. Time-over-threshold readout to enhance the high flux capabilities of single-photon-counting detectors

    International Nuclear Information System (INIS)

    Bergamaschi, Anna; Dinapoli, Roberto; Greiffenberg, Dominic; Henrich, Beat; Johnson, Ian; Mozzanica, Aldo; Radicci, Valeria; Schmitt, Bernd; Shi, Xintian; Stoppani, Laura

    2011-01-01

    The MYTHEN photon-counting ASIC operated in time-over-threshold mode shows an innovative approach towards the development of a detector operating with very high photon intensities while maintaining the single-photon sensitivity for synchrotron radiation experiments. The MYTHEN single-photon-counting (SPC) detector has been characterized using the time-over-threshold (ToT) readout method, i.e. measuring the time that the signal produced by the detected X-rays remains above the comparator threshold. In the following it is shown that the ToT readout preserves the sensitivity, dynamic range and capability of background suppression of the SPC mode, while enhancing the count-rate capability, which is the main limitation of state-of-the-art SPC systems

  13. Time-over-threshold readout to enhance the high flux capabilities of single-photon-counting detectors

    Energy Technology Data Exchange (ETDEWEB)

    Bergamaschi, Anna, E-mail: anna.bergamaschi@psi.ch; Dinapoli, Roberto; Greiffenberg, Dominic; Henrich, Beat; Johnson, Ian; Mozzanica, Aldo; Radicci, Valeria; Schmitt, Bernd; Shi, Xintian; Stoppani, Laura [Paul Scherrer Institut, CH-5232 Villigen (Switzerland)

    2011-11-01

    The MYTHEN photon-counting ASIC operated in time-over-threshold mode shows an innovative approach towards the development of a detector operating with very high photon intensities while maintaining the single-photon sensitivity for synchrotron radiation experiments. The MYTHEN single-photon-counting (SPC) detector has been characterized using the time-over-threshold (ToT) readout method, i.e. measuring the time that the signal produced by the detected X-rays remains above the comparator threshold. In the following it is shown that the ToT readout preserves the sensitivity, dynamic range and capability of background suppression of the SPC mode, while enhancing the count-rate capability, which is the main limitation of state-of-the-art SPC systems.

  14. Information Entropy Production of Maximum Entropy Markov Chains from Spike Trains

    Science.gov (United States)

    Cofré, Rodrigo; Maldonado, Cesar

    2018-01-01

    We consider the maximum entropy Markov chain inference approach to characterize the collective statistics of neuronal spike trains, focusing on the statistical properties of the inferred model. We review large deviations techniques useful in this context to describe properties of accuracy and convergence in terms of sampling size. We use these results to study the statistical fluctuation of correlations, distinguishability and irreversibility of maximum entropy Markov chains. We illustrate these applications using simple examples where the large deviation rate function is explicitly obtained for maximum entropy models of relevance in this field.

  15. Furosemide-caused alterations in the renogram of patients with renal disease and hypertension

    International Nuclear Information System (INIS)

    Camargo, E.E.; Papaleo Netto, M.; Dias Neto, A.L.; Carvalho, N.

    1975-01-01

    In 21 patients suffering - separately or simultaneously - from hypertension or nephropathy, the behavior of the differential effective renal plasma flow (ERPF sub(R) and ERPF sub(D)) and of the following time parameters of the renogram: time from point of injection to point of maximum count rate (T sub(max)), time from maximum count rate to point of 1/2 maximum count (T sub(1/2)), 'secretory' (T sub(S)) 'excretory' (T sub(ex)) half-lives of hippuran- 131 I are studied, before and after administration of furosemide. The diuretic markedly shortens all the time parameters studied, but does not change the differential effective renal plasma flow. Results suggest an action of the drug on the proximal tubule and also on loop of Henle. Neither the rate of the supply of the tracer to the tubular cell nor the tracer's pathway are altered by the diuretic [pt

  16. Effect of usual lead apron in decreasing dose rate in nuclear medicine department

    Energy Technology Data Exchange (ETDEWEB)

    Momennezhad, M.; Ghazikhanloo, K.; Zakavi, S.R. [Mashhad Univ. of Medical Sciences (Iran, Islamic Republic of)

    2006-07-01

    Full text of publication follows: Introduction: In a busy nuclear medicine department personnel exposure to radiation is inevitable during patient positioning and radiotracer preparation. There is controversy regarding usage of usual lead aprons with respect to penetrating gamma rays used in nuclear medicine departments as well as production of characteristic lead x-ray from aprons.This study tries to evaluate the effect of 0.5 mm lead apron on dose reduction. Methods and materials: We used three point sources usual radioisotopes used in nuclear medicine departments (99 mTc, 201 Tl and 131 I) and a single head L.F.O.V. gamma camera (S.M.V., D.S.X.) was used for counting purposes. The count rate for each source was about 20 K count/sec in air and in a brain water phantom. All point sources were placed at a distance of 3 meters, parallel to the center of the detector. The collimator was removed and imaging was performed for 1 min with and without lead apron on the detector. Lead apron covered all field of view of the detector. Each imaging was repeated 3 times and mean count was obtained for each radioisotope. The measurements were accomplished from full spectrum and specific region of spectrum such as characteristic lead x rays region (88 keV 20%) with and without apron. Result and Discussion: The measurement and comparison of count rates (count/min) for each source in different conditions (with and without apron source in air and in water phantom) showed that count rates were reduced in air about 77.3%, 84.2% and 40.8% for 99 mTc, 201 Tl and 131 I respectively. The reduction in count rates when sources placed in brain water phantom were 83.5%, 87% and 53.7% for the same isotope respectively. As the main source of radiation for personnel is from scattered photon and with respect to about 83% of count rate reduction using lead aprons for 99 mTc, it is expected that wearing lead apron significantly decrease dose rate. Conclusion: Our study showed that lead aprons

  17. Effect of usual lead apron in decreasing dose rate in nuclear medicine department

    International Nuclear Information System (INIS)

    Momennezhad, M.; Ghazikhanloo, K.; Zakavi, S.R.

    2006-01-01

    Full text of publication follows: Introduction: In a busy nuclear medicine department personnel exposure to radiation is inevitable during patient positioning and radiotracer preparation. There is controversy regarding usage of usual lead aprons with respect to penetrating gamma rays used in nuclear medicine departments as well as production of characteristic lead x-ray from aprons.This study tries to evaluate the effect of 0.5 mm lead apron on dose reduction. Methods and materials: We used three point sources usual radioisotopes used in nuclear medicine departments (99 mTc, 201 Tl and 131 I) and a single head L.F.O.V. gamma camera (S.M.V., D.S.X.) was used for counting purposes. The count rate for each source was about 20 K count/sec in air and in a brain water phantom. All point sources were placed at a distance of 3 meters, parallel to the center of the detector. The collimator was removed and imaging was performed for 1 min with and without lead apron on the detector. Lead apron covered all field of view of the detector. Each imaging was repeated 3 times and mean count was obtained for each radioisotope. The measurements were accomplished from full spectrum and specific region of spectrum such as characteristic lead x rays region (88 keV 20%) with and without apron. Result and Discussion: The measurement and comparison of count rates (count/min) for each source in different conditions (with and without apron source in air and in water phantom) showed that count rates were reduced in air about 77.3%, 84.2% and 40.8% for 99 mTc, 201 Tl and 131 I respectively. The reduction in count rates when sources placed in brain water phantom were 83.5%, 87% and 53.7% for the same isotope respectively. As the main source of radiation for personnel is from scattered photon and with respect to about 83% of count rate reduction using lead aprons for 99 mTc, it is expected that wearing lead apron significantly decrease dose rate. Conclusion: Our study showed that lead aprons

  18. Photon-counting image sensors

    CERN Document Server

    Teranishi, Nobukazu; Theuwissen, Albert; Stoppa, David; Charbon, Edoardo

    2017-01-01

    The field of photon-counting image sensors is advancing rapidly with the development of various solid-state image sensor technologies including single photon avalanche detectors (SPADs) and deep-sub-electron read noise CMOS image sensor pixels. This foundational platform technology will enable opportunities for new imaging modalities and instrumentation for science and industry, as well as new consumer applications. Papers discussing various photon-counting image sensor technologies and selected new applications are presented in this all-invited Special Issue.

  19. Increasing the collection efficiency of time-correlated single-photon counting with single-photon avalanche diodes using immersion lenses.

    Science.gov (United States)

    Pichette, Charles; Giudice, Andrea; Thibault, Simon; Bérubé-Lauzière, Yves

    2016-11-20

    Single-photon avalanche diodes (SPADs) achieving high timing resolution (≈20-50  ps) developed for time-correlated single-photon counting (TCSPC) generally have very small photosensitive areas (25-100 μm in diameter). This limits the achievable photon counting rate and signal-to-noise ratio and may lead to long counting times. This is detrimental in applications requiring several measurements, such as fluorescence lifetime imaging (FLIM) microscopy, which requires scanning, and time-domain diffuse optical tomography (TD-DOT). We show in this work that the use of an immersion lens directly affixed onto the photosensitive area of the SPAD helps alleviate this problem by allowing more light to be concentrated onto the detector. Following careful optical design and simulations, our experimental results show that it is actually possible to achieve the predicted theoretical increase in the photon counting rate (we achieve a factor of ≈4 here). This work is of high relevance in high timing resolution TCSPC with small photosensitive area detectors and should find widespread interest in FLIM and TD-DOT with SPADs.

  20. Rating Movies and Rating the Raters Who Rate Them.

    Science.gov (United States)

    Zhou, Hua; Lange, Kenneth

    2009-11-01

    The movie distribution company Netflix has generated considerable buzz in the statistics community by offering a million dollar prize for improvements to its movie rating system. Among the statisticians and computer scientists who have disclosed their techniques, the emphasis has been on machine learning approaches. This article has the modest goal of discussing a simple model for movie rating and other forms of democratic rating. Because the model involves a large number of parameters, it is nontrivial to carry out maximum likelihood estimation. Here we derive a straightforward EM algorithm from the perspective of the more general MM algorithm. The algorithm is capable of finding the global maximum on a likelihood landscape littered with inferior modes. We apply two variants of the model to a dataset from the MovieLens archive and compare their results. Our model identifies quirky raters, redefines the raw rankings, and permits imputation of missing ratings. The model is intended to stimulate discussion and development of better theory rather than to win the prize. It has the added benefit of introducing readers to some of the issues connected with analyzing high-dimensional data.

  1. A ppendicitis with Positive Ultrasound and its Relation to Total and Differential Leucocytic Count

    International Nuclear Information System (INIS)

    Saleh, G.M.; Sabek, E.A.S.; Elessiely, E.S.; Ibrahim, E.M.

    2016-01-01

    Background:The most common cause of acute surgical abdomen is acute appendicitis and the most commonly done emergency surgery is appendectomy. Although a battery of tests and different scoring methods are available for diagnosis of acute appendicitis, it is very difficult to prevent negative explorations for appendicectomy (15 - 30%). None of the tests has satisfactory Sensitivity and specificity that can be relied upon. Objectives: The aim of the present study was to evaluate the role of total and differential leukocyte count, and ultrasonography (USG) of abdomen in diagnosing acute appendicitis and reducing the rates of negative appendicectomies and to prevent complications like perforation, peritonitis, and the longer the delay between diagnosis and surgery, the more likely is the perforation leading to peritonitis. In addition, emphasis was given to whether combining the investigations for the same patient would improve the diagnostic accuracy. Materials and Methods : A total of 76 clinically diagnosed patients of acute appendicitis, posted for emergency appendicectomy were included in the study in General Surgery Department of EL-Salam Hospital. This study was conducted in the period October 2012 to September 2013. Preoperatively blood tests for, total leukocyte count, differential leukocyte count(DLC) and USG abdomen were done. All patients were subjected to histological examination postoperatively, which was taken as the gold standard. The three investigations results were correlated with histopathological examination reports to evaluate their role in diagnosis of acute appendicitis. Results: In the present study, the US has the highest sensitivity and specificity (90%, 80%) followed by Total leucocyte count (87.5%, 90%), white Neutrophil count (78.75%, 80%). Combining US and WBC count increases the sensitivity and specificity of the tests (96.25%, 80%). Conclusion: US contains important diagnostic information and hence should always be included in the

  2. Platelet counting using the Coulter electronic counter.

    Science.gov (United States)

    Eggleton, M J; Sharp, A A

    1963-03-01

    A method for counting platelets in dilutions of platelet-rich plasm using the Coulter electronic counter is described.(1) The results obtained show that such platelet counts are at least as accurate as the best methods of visual counting. The various technical difficulties encountered are discussed.

  3. Characterization of actinide targets by low solid-angle alpha particle counting

    CERN Document Server

    Denecke, B; Pauwels, J; Robouch, P; Gilliam, D M; Hodge, P; Hutchinson, J M R; Nico, J S

    1999-01-01

    Actinide samples were characterized in an interlaboratory comparison between IRMM and NIST, including alpha-particle counting at defined low solid angle and counting in a 2 pi proportional gas counter. For this comparison, nine sup 2 sup 3 sup 3 UF sub 4 samples with high uniformity in the layer thickness were prepared at IRMM by deposition under vacuum. Polished silicon wafers were used as source substrates, and these were rotated during the deposition using a planetary rotation system. The estimated uncertainties for the defined low solid-angle methods were about 0.1% at both NIST and IRMM. The agreement of reported alpha-particle emission rates in the energy range 2.5-5.09 MeV was better than or equal to 0.02% for the defined solid-angle methods. When comparing total alpha-particle emission rates over the larger energy range 0-9 MeV (which includes all emissions from the daughter nuclides and the impurities), the agreement of the defined solid-angle methods was better than or equal to 0.05%. The 2 pi propo...

  4. Determination of potassium in feldspars by beta counting using a GM multicounter system

    Energy Technology Data Exchange (ETDEWEB)

    Boetter-Jensen, L; Mejdahl, V

    1985-01-01

    A gas flow multicounter system developed at Risoe National laboratory for low-level beta counting applications was adopted for measuring potassium in feldspars extracted from archaeological and geological materials, in connection with TL dating work. A guard counter reduces the background by using an anticoincidence technique. Calibration was accomplished by measuring KCl and a feldspar standard. A potassium content of 1% yields a net beta count rate (background subtracted) of about 40 c.p.h.; the background is around 15 c.p.h. The precision in repeated measurements is better than 2% and the estimated accuracy is about 3%.

  5. Grade Assignment by Ki-67 Proliferative Index, Mitotic Count, and Phosphohistone H3 Count in Surgically Resected Gastrointestinal and Pancreatic Neuroendocrine Tumors.

    Science.gov (United States)

    Murphy, Claire E; McCormick, Kinsey A; Shankaran, Veena; Reddi, Deepti M; Swanson, Paul E; Upton, Melissa P; Papanicolau-Sengos, Antonios; Khor, Sara; Westerhoff, Maria

    The aim of this study was to evaluate the concordance in grade assignment for gastroenteropancreatic neuroendocrine tumors using mitotic count (MC), Ki-67 proliferative index (KPI), and phosphohistone H3 count (PHH3C). Resected gastroenteropancreatic neuroendocrine tumors were graded based on MC, KPI, and PHH3C. Concordance was determined using a weighted κ statistic. Median survival across each grade category was determined using Kaplan-Meier methods. Of the 110 patients, the majority had gastrointestinal primaries and grade 1 or 2 tumors. Rates of discordance in grade assignment were 29% of cases for KPI versus MC (κW = 0.26), 32% for PHH3C versus MC (κW = 0.34), and 32% for PHH3C versus KPI (κW = 0.37). There was fair agreement between grading by KPI and MC. Relative to grade by KPI and MC, PHH3C tended to upgrade tumors. The proportion alive at 3 and 5 years was not significantly different for patients with grade 1 versus grade 2 tumors. The concordance between KPI and MC was fair. Phosphohistone H3 count tended to upgrade tumors using the cutoffs established by MC. Grade 1 and grade 2 tumors were associated with similar survival regardless of grading method. The overall relevance of the current cutoff values used in grading neuroendocrine tumors may need to be revisited.

  6. Comparative evaluation of platelet count and antimicrobial efficacy of injectable platelet-rich fibrin with other platelet concentrates: An in vitro study

    Directory of Open Access Journals (Sweden)

    Prerna Ashok Karde

    2017-01-01

    Full Text Available Background: Platelet concentrates are used in various medical procedures to promote soft- and hard-tissue regeneration. In recent times, their antimicrobial efficacy is also explored. However, various platelet concentrates have evolved which differ in the centrifugation protocols. One such recently introduced platelet concentrate is injectable platelet-rich fibrin (i-PRF concentrate. Hence, the aim was to evaluate the antimicrobial property, and platelet count of i-PRF in comparison to other platelet concentrates, i.e., PRF, platelet-rich plasma (PRP, and control (whole blood. Materials and Methods: Blood samples were obtained from 10 chronic generalized marginal gingivitis patients. Platelet concentrates were prepared using standardized centrifugation protocol. Platelet count was evaluated by manual counting method using smear preparation of each sample. Subsequently, antimicrobial activity against oral bacteria was examined on blood agar using disc diffusion method to quantify the inhibitory effects. Results: Statistical significance was analyzed by one-way analysis of variance (ANOVA. P 0.05. i-PRF showed statistically significant difference (P < 0.001 in platelet count when compared to control. It was also significant when compared to PRP (P < 0.01, PRF (P < 0.001. Conclusion: i-PRF has maximum antimicrobial efficacy and higher platelet count in comparison to other platelet concentrates, thereby indicating to have a better regenerative potential then others.

  7. A low background-rate detector for ions in the 5 to 50 keV energy range to be used for radioisotope dating with a small cyclotron

    International Nuclear Information System (INIS)

    Friedman, P.G.

    1986-01-01

    Accelerator mass spectrometry in tandem Van de Graaff accelerators has proven successful for radioisotope dating small samples. We are developing a 20 cm diameter 30 to 40 keV cyclotron dedicated to high-sensitivity radioisotope dating, initially for 14 C. At this energy, range and dE/dx methods of particle identification are impossible. Thus arises the difficult problem of reliably detecting 30 to 40 keV 14 C at 10 -2 counts/sec in the high background environment of the cyclotron, where lower energy ions, electrons, and photons bombard the detector at much higher rates. We have developed and tested an inexpensive, generally useful ion detector that allows dark-count rates below 10 -4 counts/sec and excellent background suppression. With the cyclotron tuned near the 13 CH background peak, to the frequency for 14 C, the detector suppresses the background to 6 x 10 -4 counts/sec. For each 14 C ion the detectors grazing-incidence Al 2 O 3 conversion dynode emits about 20 secondary electrons, which are independently multiplied in separate pores of a microchannel plate. The output signal is proportional to the number of secondary electrons, allowing pulse-height discrimination of background. We have successfully tested the detector with positive 12 C, 23 Na, 39 K, 41 K, 85 Rb, 87 Rb, and 133 Cs at 5 to 40 keV, and with 36 keV negative 12 C and 13 CH. It should detect ions and neutrals of all species, at energies above 5 keV, with good efficiency and excellent background discrimination. Counting efficiency and background discrimination improve with higher ion energy. The detector can be operated at least up to 2 x 10 -7 Torr and be repeatedly exposed to air. The maximum rate is 10/sup 6.4/ ions/sec in pulse counting mode and 10/sup 9.7/ ions/sec in current integrating mode

  8. Resonance ionization spectroscopy: Counting noble gas atoms

    International Nuclear Information System (INIS)

    Hurst, G.S.; Payne, M.G.; Chen, C.H.; Willis, R.D.; Lehmann, B.E.; Kramer, S.D.

    1981-01-01

    The purpose of this paper is to describe new work on the counting of noble gas atoms, using lasers for the selective ionization and detectors for counting individual particles (electrons or positive ions). When positive ions are counted, various kinds of mass analyzers (magnetic, quadrupole, or time-of-flight) can be incorporated to provide A selectivity. We show that a variety of interesting and important applications can be made with atom-counting techniques which are both atomic number (Z) and mass number (A) selective. (orig./FKS)

  9. Design Study of an Incinerator Ash Conveyor Counting System - 13323

    International Nuclear Information System (INIS)

    Jaederstroem, Henrik; Bronson, Frazier

    2013-01-01

    A design study has been performed for a system that should measure the Cs-137 activity in ash from an incinerator. Radioactive ash, expected to consist of both Cs-134 and Cs-137, will be transported on a conveyor belt at 0.1 m/s. The objective of the counting system is to determine the Cs-137 activity and direct the ash to the correct stream after a diverter. The decision levels are ranging from 8000 to 400000 Bq/kg and the decision error should be as low as possible. The decision error depends on the total measurement uncertainty which depends on the counting statistics and the uncertainty in the efficiency of the geometry. For the low activity decision it is necessary to know the efficiency to be able to determine if the signal from the Cs-137 is above the minimum detectable activity and that it generates enough counts to reach the desired precision. For the higher activity decision the uncertainty of the efficiency needs to be understood to minimize decision errors. The total efficiency of the detector is needed to be able to determine if the detector will be able operate at the count rate at the highest expected activity. The design study that is presented in this paper describes how the objectives of the monitoring systems were obtained, the choice of detector was made and how ISOCS (In Situ Object Counting System) mathematical modeling was used to calculate the efficiency. The ISOCS uncertainty estimator (IUE) was used to determine which parameters of the ash was important to know accurately in order to minimize the uncertainty of the efficiency. The examined parameters include the height of the ash on the conveyor belt, the matrix composition and density and relative efficiency of the detector. (authors)

  10. Design Study of an Incinerator Ash Conveyor Counting System - 13323

    Energy Technology Data Exchange (ETDEWEB)

    Jaederstroem, Henrik; Bronson, Frazier [Canberra Industries Inc., 800 Research Parkway Meriden CT 06450 (United States)

    2013-07-01

    A design study has been performed for a system that should measure the Cs-137 activity in ash from an incinerator. Radioactive ash, expected to consist of both Cs-134 and Cs-137, will be transported on a conveyor belt at 0.1 m/s. The objective of the counting system is to determine the Cs-137 activity and direct the ash to the correct stream after a diverter. The decision levels are ranging from 8000 to 400000 Bq/kg and the decision error should be as low as possible. The decision error depends on the total measurement uncertainty which depends on the counting statistics and the uncertainty in the efficiency of the geometry. For the low activity decision it is necessary to know the efficiency to be able to determine if the signal from the Cs-137 is above the minimum detectable activity and that it generates enough counts to reach the desired precision. For the higher activity decision the uncertainty of the efficiency needs to be understood to minimize decision errors. The total efficiency of the detector is needed to be able to determine if the detector will be able operate at the count rate at the highest expected activity. The design study that is presented in this paper describes how the objectives of the monitoring systems were obtained, the choice of detector was made and how ISOCS (In Situ Object Counting System) mathematical modeling was used to calculate the efficiency. The ISOCS uncertainty estimator (IUE) was used to determine which parameters of the ash was important to know accurately in order to minimize the uncertainty of the efficiency. The examined parameters include the height of the ash on the conveyor belt, the matrix composition and density and relative efficiency of the detector. (authors)

  11. ROC [Receiver Operating Characteristics] study of maximum likelihood estimator human brain image reconstructions in PET [Positron Emission Tomography] clinical practice

    International Nuclear Information System (INIS)

    Llacer, J.; Veklerov, E.; Nolan, D.; Grafton, S.T.; Mazziotta, J.C.; Hawkins, R.A.; Hoh, C.K.; Hoffman, E.J.

    1990-10-01

    This paper will report on the progress to date in carrying out Receiver Operating Characteristics (ROC) studies comparing Maximum Likelihood Estimator (MLE) and Filtered Backprojection (FBP) reconstructions of normal and abnormal human brain PET data in a clinical setting. A previous statistical study of reconstructions of the Hoffman brain phantom with real data indicated that the pixel-to-pixel standard deviation in feasible MLE images is approximately proportional to the square root of the number of counts in a region, as opposed to a standard deviation which is high and largely independent of the number of counts in FBP. A preliminary ROC study carried out with 10 non-medical observers performing a relatively simple detectability task indicates that, for the majority of observers, lower standard deviation translates itself into a statistically significant detectability advantage in MLE reconstructions. The initial results of ongoing tests with four experienced neurologists/nuclear medicine physicians are presented. Normal cases of 18 F -- fluorodeoxyglucose (FDG) cerebral metabolism studies and abnormal cases in which a variety of lesions have been introduced into normal data sets have been evaluated. We report on the results of reading the reconstructions of 90 data sets, each corresponding to a single brain slice. It has become apparent that the design of the study based on reading single brain slices is too insensitive and we propose a variation based on reading three consecutive slices at a time, rating only the center slice. 9 refs., 2 figs., 1 tab

  12. Counting and Surveying Homeless Youth: Recommendations from YouthCount 2.0!, a Community-Academic Partnership.

    Science.gov (United States)

    Narendorf, Sarah C; Santa Maria, Diane M; Ha, Yoonsook; Cooper, Jenna; Schieszler, Christine

    2016-12-01

    Communities across the United States are increasing efforts to find and count homeless youth. This paper presents findings and lessons learned from a community/academic partnership to count homeless youth and conduct an in depth research survey focused on the health needs of this population. Over a 4 week recruitment period, 632 youth were counted and 420 surveyed. Methodological successes included an extended counting period, broader inclusion criteria to capture those in unstable housing, use of student volunteers in health training programs, recruiting from magnet events for high risk youth, and partnering with community agencies to disseminate findings. Strategies that did not facilitate recruitment included respondent driven sampling, street canvassing beyond known hotspots, and having community agencies lead data collection. Surveying was successful in gathering data on reasons for homelessness, history in public systems of care, mental health history and needs, sexual risk behaviors, health status, and substance use. Youth were successfully surveyed across housing types including shelters or transitional housing (n = 205), those in unstable housing such as doubled up with friends or acquaintances (n = 75), and those who were literally on the streets or living in a place not meant for human habitation (n = 140). Most youth completed the self-report survey and provided detailed information about risk behaviors. Recommendations to combine research data collection with counting are presented.

  13. Triton X-100 as a complete liquid scintillation cocktail for counting aqueous solutions and ionic nutrient salts

    International Nuclear Information System (INIS)

    Reed, D.W.

    1984-01-01

    Triton X-100, used alone, was found to act as a complete liquid scintillation cocktail. Triton X-100 acted as a scintillator and the effect was not due to Cerenkov radiation. A variety of other commercially available surfactants also acted as scintillators, but with different levels of efficiency. Triton X-100/water combinations were suitable for counting aqueous solutions of 33 P and 86 Rb and the count rate was stable over extended periods of time. Triton X-100/toluene combinations also yielded high counting efficiencies. Triton X-100 was more sensitive to quenching than standard cocktails containing fluors. (author)

  14. A yearly spraying of olive mill wastewater on agricultural soil over six successive years: impact of different application rates on olive production, phenolic compounds, phytotoxicity and microbial counts.

    Science.gov (United States)

    Magdich, Salwa; Jarboui, Raja; Rouina, Béchir Ben; Boukhris, Makki; Ammar, Emna

    2012-07-15

    Olive mill wastewater (OMW) spraying effects onto olive-tree fields were investigated. Three OMW levels (50, 100 and 200 m(3)ha(-1)year(-1)) were applied over six successive years. Olive-crop yields, phenolic compounds progress, phytotoxicity and microbial counts were studied at different soil depths. Olive yield showed improvements with OMW level applied. Soil polyphenolic content increased progressively in relation to OMW levels in all the investigated layers. However, no significant difference was noted in lowest treatment rate compared to the control field. In the soil upper-layers (0-40 cm), five phenolic compounds were identified over six consecutive years of OMW-spraying. In all the soil-layers, the radish germination index exceeded 85%. However, tomato germination test values decreased with the applied OMW amount. For all treatments, microbial counts increased with OMW quantities and spraying frequency. Matrix correlation showed a strong relationship between soil polyphenol content and microorganisms, and a negative one to tomato germination index. Copyright © 2012 Elsevier B.V. All rights reserved.

  15. CalCOFI Larvae Counts Positive Tows

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Fish larvae counts and standardized counts for eggs captured in CalCOFI icthyoplankton nets (primarily vertical [Calvet or Pairovet], oblique [bongo or ring nets],...

  16. Set of counts by scintillations for atmospheric samplings

    International Nuclear Information System (INIS)

    Appriou, D.; Doury, A.

    1962-01-01

    The author reports the development of a scintillation-based counting assembly with the following characteristics: a photo-multiplier with a wide photo-cathode, a thin plastic scintillator for the counting of beta + alpha (and possibility of mounting an alpha scintillator), a relatively small own motion with respect to activities to be counted, a weakly varying efficiency. The authors discuss the counting objective, present equipment tests (counter, proportional amplifier and pre-amplifier, input drawer). They describe the apparatus operation, discuss the selection of scintillators, report the study of the own movement (electron-based background noise, total background noise, background noise reduction), discuss counts (influence of the external source, sensitivity to alpha radiations, counting homogeneity, minimum detectable activity) and efficiencies

  17. Whales from space: counting southern right whales by satellite.

    Science.gov (United States)

    Fretwell, Peter T; Staniland, Iain J; Forcada, Jaume

    2014-01-01

    We describe a method of identifying and counting whales using very high resolution satellite imagery through the example of southern right whales breeding in part of the Golfo Nuevo, Península Valdés in Argentina. Southern right whales have been extensively hunted over the last 300 years and although numbers have recovered from near extinction in the early 20(th) century, current populations are fragmented and are estimated at only a small fraction of pre-hunting total. Recent extreme right whale calf mortality events at Península Valdés, which constitutes the largest single population, have raised fresh concern for the future of the species. The WorldView2 satellite has a maximum 50 cm resolution and a water penetrating coastal band in the far-blue part of the spectrum that allows it to see deeper into the water column. Using an image covering 113 km², we identified 55 probable whales and 23 other features that are possibly whales, with a further 13 objects that are only detected by the coastal band. Comparison of a number of classification techniques, to automatically detect whale-like objects, showed that a simple thresholding technique of the panchromatic and coastal band delivered the best results. This is the first successful study using satellite imagery to count whales; a pragmatic, transferable method using this rapidly advancing technology that has major implications for future surveys of cetacean populations.

  18. Track counting in radon dosimetry

    International Nuclear Information System (INIS)

    Fesenbeck, Ingo; Koehler, Bernd; Reichert, Klaus-Martin

    2013-01-01

    The newly developed, computer-controlled track counting system is capable of imaging and analyzing the entire area of nuclear track detectors. The high optical resolution allows a new analysis approach for the process of automated counting using digital image processing technologies. This way, higher exposed detectors can be evaluated reliably by an automated process as well. (orig.)

  19. Upper limits of the photon fluence rate on CT detectors: Case study on a commercial scanner

    Energy Technology Data Exchange (ETDEWEB)

    Persson, Mats, E-mail: mats.persson@mi.physics.kth.se; Bornefalk, Hans; Danielsson, Mats [Department of Physics, Royal Institute of Technology, Stockholm SE-10691 (Sweden); Bujila, Robert; Nowik, Patrik; Andersson, Henrik [Unit of X-ray Physics, Section of Imaging Physics Solna, Department of Medical Physics, Karolinska University Hospital, Stockholm SE-17176 (Sweden); Kull, Love [Medical Radiation Physics, Sunderby Hospital, Luleå SE-97180 (Sweden); Andersson, Jonas [Department of Radiation Sciences, Radiation Physics, Umeå University, Umeå SE-90185 (Sweden)

    2016-07-15

    Purpose: The highest photon fluence rate that a computed tomography (CT) detector must be able to measure is an important parameter. The authors calculate the maximum transmitted fluence rate in a commercial CT scanner as a function of patient size for standard head, chest, and abdomen protocols. Methods: The authors scanned an anthropomorphic phantom (Kyoto Kagaku PBU-60) with the reference CT protocols provided by AAPM on a GE LightSpeed VCT scanner and noted the tube current applied with the tube current modulation (TCM) system. By rescaling this tube current using published measurements on the tube current modulation of a GE scanner [N. Keat, “CT scanner automatic exposure control systems,” MHRA Evaluation Report 05016, ImPACT, London, UK, 2005], the authors could estimate the tube current that these protocols would have resulted in for other patient sizes. An ECG gated chest protocol was also simulated. Using measured dose rate profiles along the bowtie filters, the authors simulated imaging of anonymized patient images with a range of sizes on a GE VCT scanner and calculated the maximum transmitted fluence rate. In addition, the 99th and the 95th percentiles of the transmitted fluence rate distribution behind the patient are calculated and the effect of omitting projection lines passing just below the skin line is investigated. Results: The highest transmitted fluence rates on the detector for the AAPM reference protocols with centered patients are found for head images and for intermediate-sized chest images, both with a maximum of 3.4 ⋅ 10{sup 8} mm{sup −2} s{sup −1}, at 949 mm distance from the source. Miscentering the head by 50 mm downward increases the maximum transmitted fluence rate to 5.7 ⋅ 10{sup 8} mm{sup −2} s{sup −1}. The ECG gated chest protocol gives fluence rates up to 2.3 ⋅ 10{sup 8} − 3.6 ⋅ 10{sup 8} mm{sup −2} s{sup −1} depending on miscentering. Conclusions: The fluence rate on a CT detector reaches 3 ⋅ 10{sup 8

  20. CalCOFI Egg Counts Positive Tows

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Fish egg counts and standardized counts for eggs captured in CalCOFI icthyoplankton nets (primarily vertical [Calvet or Pairovet], oblique [bongo or ring nets], and...