Sample records for left-sided pacemaker specification

  1. Importance of a persisting left vena cava superior in pacemaker therapy

    International Nuclear Information System (INIS)

    Marosi, G.; Sagi, J.; Pokorny, L.; Simon, Z.; Orvostudomangi Egyetem, Szeged

    1982-01-01

    Basing on the description of a specific case and a review of literature, the article discusses the relationship between persisting left vena cava superior and pacemaker therapy. The article covers the technical problems caused by the vascular anomaly, the complications which can be expected, the possibilities in respect of recognising the vascular anomaly, as well as the methods for coping with the special situation prevailing in a particular case. (orig.) [de

  2. Neck Pain One Week after Pacemaker Generator Replacement.

    Science.gov (United States)

    Graham, Ross F; Wightman, John M

    2015-07-01

    The incidence of cardiac pacemaker implantation has risen markedly in the past three decades, making awareness of possible postprocedural complications critical to the emergency physician. This case is the first documented instance of internal jugular (IJ) deep vein thrombosis (DVT) from an uncomplicated pacemaker generator replacement. A patient presented to an Emergency Department with a 2-day history of mild left temporal headache migrating to his left neck. The patient did not volunteer this information, but review of systems revealed a temporary transvenous pacemaker inserted through the right IJ vein 1 week previously during a routine exchange of a left-sided cardiac pacemaker generator. Manipulation of the existing pacemaker wires entering the left subclavian vein was minimal. Computed tomographic angiography of the neck demonstrated near-complete thrombotic occlusion of the entire length of his left IJ vein. This required hospital admission for observation and treatment with anticoagulation. WHY SHOULD AN EMERGENCY PHYSICIAN BE AWARE OF THIS?: DVT, with thrombotic extension into adjacent vessels anywhere along the course of pacemaker wires, should be considered by the emergency provider in the evaluation of head, neck, or upper extremity symptoms after recent or remote implantation or manipulation of a transvenous cardiac pacemaker, including generator replacement. Failure to identify and treat appropriately could result in significant morbidity and mortality from airway edema, septic thrombophlebitis, superior vena cava syndrome, superior sagittal sinus thrombosis, or pulmonary embolism. Published by Elsevier Inc.

  3. Pacemaker syndrome with sub-acute left ventricular systolic dysfunction in a patient with a dual-chamber pacemaker: consequence of lead switch at the header.

    Science.gov (United States)

    Khurwolah, Mohammad Reeaze; Vezi, Brian Zwelethini

    In the daily practice of pacemaker insertion, the occurrence of atrial and ventricular lead switch at the pacemaker box header is a rare and unintentional phenomenon, with less than five cases reported in the literature. The lead switch may have dire consequences, depending on the indication for the pacemaker. One of these consequences is pacemaker syndrome, in which the normal sequence of atrial and ventricular activation is impaired, leading to sub-optimal ventricular filling and cardiac output. It is important for the attending physician to recognise any worsening of symptoms in a patient who has recently had a permanent pacemaker inserted. In the case of a dual-chamber pacemaker, switching of the atrial and ventricular leads at the pacemaker box header should be strongly suspected. We present an unusual case of pacemaker syndrome and right ventricular-only pacinginduced left ventricular systolic dysfunction in a patient with a dual-chamber pacemaker.

  4. Risk of pacemaker implantation subsequent to radiotherapy for early-stage breast cancer in Denmark, 1982-2005

    DEFF Research Database (Denmark)

    Rehammar, Jens Christian; Videbaek, L.; Brock Johansen, J.

    2015-01-01

    . Objectives: The aim of this study was to examine the risk of severe conduction abnormalities evaluated by implantation of a pacemaker, subsequent to breast cancer radiotherapy. Methods: From the database of the Danish Breast Cancer Collaborative Group, we identified women treated with radiotherapy for early......-stage breast cancer in Denmark from 1982 to 2005. By record linkage to the Danish Pacemaker and ICD Registry information was retrieved on pacemaker implants subsequent to radiotherapy. Rate ratios (RR) of pacemaker implantation for left versus right sided breast cancer were calculated. Results: Among 18......,308 women treated with radiotherapy for early-stage breast cancer, 179 women had a pacemaker implanted subsequent to radiotherapy, 90 in 9,315 left sided and 89 in 8,993 right sided breast cancers. The unadjusted RR was 1.02 (0.76-1.36 95% CI, p=0.91) and the RR adjusted for year, age and time since...

  5. Superior transseptal approach to mitral valve is associated with a higher need for pacemaker implantation than the left atrial approach

    DEFF Research Database (Denmark)

    Lukac, Peter; Hjortdal, Vibeke E; Pedersen, Anders K

    2006-01-01

    retrospectively evaluated. The surgeons used either the superior transseptal (group A) or left atrial approach (group B). The risk of pacemaker implantation associated with the superior transseptal approach as compared with the left atrial approach was estimated using the multivariate Cox regression analysis...... to adjust for possible confounders. RESULTS: We included 577 patients, 150 in group A and 427 in group B. Forty-four patients had a pacemaker implanted after the surgery; 17 in group A and 27 in group B (p = 0.010). The superior transseptal approach was an independent risk factor of pacemaker implantation...... in multivariate analysis (hazard ratio 2.2 [1.2 to 4.1], p = 0.014). Nineteen patients had a pacemaker implanted because of sinus node dysfunction; 9 in group A and 10 in group B (p = 0.017). Group A was an independent predictor of pacemaker implantation because of sinus node dysfunction in bivariate analyses...

  6. Biochemical and biomechanical properties of the pacemaking sinoatrial node extracellular matrix are distinct from contractile left ventricular matrix.

    Directory of Open Access Journals (Sweden)

    Jessica M Gluck

    Full Text Available Extracellular matrix plays a role in differentiation and phenotype development of its resident cells. Although cardiac extracellular matrix from the contractile tissues has been studied and utilized in tissue engineering, extracellular matrix properties of the pacemaking sinoatrial node are largely unknown. In this study, the biomechanical properties and biochemical composition and distribution of extracellular matrix in the sinoatrial node were investigated relative to the left ventricle. Extracellular matrix of the sinoatrial node was found to be overall stiffer than that of the left ventricle and highly heterogeneous with interstitial regions composed of predominantly fibrillar collagens and rich in elastin. The extracellular matrix protein distribution suggests that resident pacemaking cardiomyocytes are enclosed in fibrillar collagens that can withstand greater tensile strength while the surrounding elastin-rich regions may undergo deformation to reduce the mechanical strain in these cells. Moreover, basement membrane-associated adhesion proteins that are ligands for integrins were of low abundance in the sinoatrial node, which may decrease force transduction in the pacemaking cardiomyocytes. In contrast to extracellular matrix of the left ventricle, extracellular matrix of the sinoatrial node may reduce mechanical strain and force transduction in pacemaking cardiomyocytes. These findings provide the criteria for a suitable matrix scaffold for engineering biopacemakers.

  7. Does concomitant tricuspid annuloplasty increase perioperative mortality and morbidity when correcting left-sided valve disease?

    Science.gov (United States)

    Zhu, Tie-Yuan; Wang, Jian-Gang; Meng, Xu

    2015-01-01

    A best evidence topic in adult valvular surgery was written according to a structured protocol. The question addressed was 'Does concomitant tricuspid annuloplasty increase the perioperative mortality and morbidity when correcting left-sided valve disease?' A total of 561 papers were found using the reported search, of which 12 presented the best evidence to answer the clinical question. The authors, country, journal, date of publication, patient group studied, study type, relevant outcomes and results of these papers are tabulated. Among these 12 papers, there were nine retrospective studies, two cohort studies and one randomized controlled trial (RCT). Overall, additional tricuspid valve (TV) repair takes more time during operations, particularly with a ring annuloplasty method. The mean aortic cross-clamping times were 57-83 min without associated tricuspid repair and 62-100 min with, and cardiopulmonary bypass times without and with repair were 82-124 and 90-174 min, respectively. A study of 624 patients who had undergone isolated mitral valve (MV) surgery and MV surgery plus TV repair showed more female and atrial fibrillation patients in the tricuspid valve plasty (TVP) group, but no increase in the 30-day mortality was found. One RCT, presenting similar patient baseline characteristics, also found no difference in the hospital mortality rates between the TVP group and the non-TVP group. Another 10 studies also demonstrated no statistically significant differences in perioperative mortality. In a cohort study of 311 patients undergoing MV repair with or without tricuspid annuloplasty, postoperative complications, such as bleeding, stroke, pacemaker, haemofiltration and myocardial infarction, all showed no statistically significant differences in the two groups. One study retrospectively analysed a large number of patients undergoing either isolated left-sided valve surgery or a concomitant TV repair, and there were no statistically significant differences

  8. Late deterioration of left ventricular function after right ventricular pacemaker implantation.

    Science.gov (United States)

    Bellmann, Barbara; Muntean, Bogdan G; Lin, Tina; Gemein, Christopher; Schmitz, Kathrin; Schauerte, Patrick

    2016-09-01

    Right ventricular (RV) pacing induces a left bundle branch block pattern on ECG and may promote heart failure. Patients with dual chamber pacemakers (DCPs) who present with progressive reduction in left ventricular ejection fraction (LVEF) secondary to RV pacing are candidates for cardiac resynchronization therapy (CRT). This study analyzes whether upgrading DCP to CRT with the additional implantation of a left ventricular (LV) lead improves LV function in patients with reduced LVEF following DCP implantation. Twenty-two patients (13 males) implanted with DCPs and a high RV pacing percentage (>90%) were evaluated in term of new-onset heart failure symptoms. The patients were enrolled in this retrospective single-center study after obvious causes for a reduced LVEF were excluded with echocardiography and coronary angiography. In all patients, DCPs were then upgraded to biventricular devices. LVEF was analyzed with a two-sided t-test. QRS duration and brain natriuretic peptide (BNP) levels were analyzed with the unpaired t-test. LVEF declined after DCP implantation from 54±10% to 31±7%, and the mean QRS duration was 161±20 ms during RV pacing. NT-pro BNP levels were elevated (3365±11436 pmol/L). After upgrading to a biventricular device, a biventricular pacing percentage of 98.1±2% was achieved. QRS duration decreased to 108±16 ms and 106±20 ms after 1 and 6 months, respectively. There was a significant increase in LVEF to 38±8% and 41±11% and a decrease in NT-pro BNP levels to 3088±2326 pmol/L and 1860±1838 pmol/L at 1 and 6 months, respectively. Upgrading to CRT may be beneficial in patients with DCPs and heart failure induced by a high RV pacing percentage.

  9. [Analysis of pacemaker ECGs].

    Science.gov (United States)

    Israel, Carsten W; Ekosso-Ejangue, Lucy; Sheta, Mohamed-Karim

    2015-09-01

    The key to a successful analysis of a pacemaker electrocardiogram (ECG) is the application of the systematic approach used for any other ECG without a pacemaker: analysis of (1) basic rhythm and rate, (2) QRS axis, (3) PQ, QRS and QT intervals, (4) morphology of P waves, QRS, ST segments and T(U) waves and (5) the presence of arrhythmias. If only the most obvious abnormality of a pacemaker ECG is considered, wrong conclusions can easily be drawn. If a systematic approach is skipped it may be overlooked that e.g. atrial pacing is ineffective, the left ventricle is paced instead of the right ventricle, pacing competes with intrinsic conduction or that the atrioventricular (AV) conduction time is programmed too long. Apart from this analysis, a pacemaker ECG which is not clear should be checked for the presence of arrhythmias (e.g. atrial fibrillation, atrial flutter, junctional escape rhythm and endless loop tachycardia), pacemaker malfunction (e.g. atrial or ventricular undersensing or oversensing, atrial or ventricular loss of capture) and activity of specific pacing algorithms, such as automatic mode switching, rate adaptation, AV delay modifying algorithms, reaction to premature ventricular contractions (PVC), safety window pacing, hysteresis and noise mode. A systematic analysis of the pacemaker ECG almost always allows a probable diagnosis of arrhythmias and malfunctions to be made, which can be confirmed by pacemaker control and can often be corrected at the touch of the right button to the patient's benefit.

  10. Pulmonary hypertension associated with left-sided heart disease.

    Science.gov (United States)

    Maeder, Micha Tobias; Schoch, Otto D; Kleiner, Rebekka; Joerg, Lucas; Weilenmann, Daniel; Swiss Society For Pulmonary Hypertension

    2017-01-19

    Pulmonary hypertension associated with left-sided heart disease (PH-LHD) is the most common type of pulmonary hypertension. In patients with left-sided heart disease, the presence of pulmonary hypertension is typically a marker of more advanced disease, more severe symptoms, and worse prognosis. In contrast to pulmonary arterial hypertension, PH-LHD is characterised by an elevated pulmonary artery wedge pressure (postcapillary pulmonary hypertension) without or with an additional precapillary component (isolated postcapillary versus combined postcapillary and precapillary pulmonary hypertension). Transthoracic echocardiography is the primary nonin-vasive imaging tool to estimate the probability of pulmonary hypertension and to establish a working diagnosis on the mechanism of pulmonary hyperten-sion. However, right heart catheterisation is always required if significant pulmonary hypertension is sus-pected and exact knowledge of the haemodynamic constellation is necessary. The haemodynamic con-stellation (mean pulmonary artery pressure, mean pulmonary artery wedge pressure, left ventricular end-diastolic pressure) in combination with clinical infor-mation and imaging findings (mainly echocardiog-raphy, coronary angiography and cardiac magnetic resonance imaging) will usually allow the exact mech-anism underlying PH-LHD to be defined, which is a prerequisite for appropriate treatment. The general principle for the management of PH-LHD is to treat the underlying left-sided heart disease in an optimal man-ner using drugs and/or interventional or surgical ther-apy. There is currently no established indication for pulmonary arterial hypertension-specific therapies in PH-LHD, and specific therapies may even cause harm in patients with PH-LHD.

  11. Anosognosia for hemiparesis after left-sided stroke.

    Science.gov (United States)

    Baier, Bernhard; Vucurevic, Goran; Müller-Forell, Wibke; Glassl, Oliver; Geber, Christian; Dieterich, Marianne; Karnath, Hans-Otto

    2014-12-01

    In patients with left-sided lesions, anosognosia for hemiparesis (AHP) seems to be a rare phenomenon. It has been discussed whether this rareness might be due to an inevitable bias due to language dysfunction and whether the left hemisphere's role for our self-awareness of motor actions thus is underestimated. By applying functional magnetic resonance imaging (fMRI) we examined whether patients with AHP following a left hemisphere stroke show a regular, left-sided or a reversed, right-sided lateralization of language functions. Only the former observation would argue for an original role of the left hemisphere in self-awareness about limb function. In a consecutive series of 44 acute left-sided stroke patients, only one patient (=2%) was identified showing AHP. In this case, we could verify by using fMRI that lateralization of AHP and spatial neglect on the one hand and of language functions on the other hand were reversed. The present single case observation thus argues against an original role of the left hemisphere in self-awareness about limb function. We discuss the data in the context of previous observations in the literature. Copyright © 2014 Elsevier Ltd. All rights reserved.

  12. Incidence of Important Hemobilia Following Transhepatic Biliary Drainage: Left-Sided Versus Right-Sided Approaches

    International Nuclear Information System (INIS)

    Rivera-Sanfeliz, G. M.; Assar, O. S. A.; LaBerge, J. M.; Wilson, M. W.; Gordon, R. L.; Ring, E. J.; Kerlan, R. K. Jr.

    2004-01-01

    Our purpose here is to describe our experience with important hemobilia following PTBD and to determine whether left-sided percutaneous transhepatic biliary drainage (PTBD) is associated with an increased incidence of important hemobilia compared to right-sided drainages. We reviewed 346 transhepatic biliary drainages over a four-year period and identified eight patients (2.3%) with important hemobilia requiring transcatheter embolization. The charts and radiographic files of these patients were reviewed. The side of the PTBD (left versus right), and the order of the biliary ductal branch entered (first, second, or third) were recorded. Of the 346 PTBDs, 269 were right-sided and 77 were left-sided. Of the eight cases of important hemobilia requiring transcatheter embolization, four followed right-sided and four followed left-sided PTBD, corresponding to a bleeding incidence of 1.5% (4/269) for right PTBD and 5.2% (4/77) for left PTBD. The higher incidence of hemobilia associated with left-sided PTBD approached, but did not reach the threshold of statistical significance (p = 0.077). In six of the eight patients requiring transcatheter embolization, first or second order biliary branches were accessed by catheter for PTBD. All patients with left-sided bleeding had first or proximal second order branches accessed by biliary drainage catheters. In conclusion, a higher incidence of hemobilia followed left- versus right-sided PTBD in this study, but the increased incidence did not reach statistical significance

  13. The pacemaker-twiddler's syndrome: an infrequent cause of pacemaker failure.

    Science.gov (United States)

    Salahuddin, Mohammad; Cader, Fathima Aaysha; Nasrin, Sahela; Chowdhury, Mashhud Zia

    2016-01-20

    The pacemaker-twiddler's syndrome is an uncommon cause of pacemaker malfunction. It occurs due to unintentional or deliberate manipulation of the pacemaker pulse generator within its skin pocket by the patient. This causes coiling of the lead and its dislodgement, resulting in failure of ventricular pacing. More commonly reported among elderly females with impaired cognition, the phenomenon usually occurs in the first year following pacemaker implantation. Treatment involves repositioning of the dislodged leads and suture fixation of the lead and pulse generator within its pocket. An 87 year old Bangladeshi lady who underwent a single chamber ventricular pacemaker (VVI mode: i.e. ventricle paced, ventricle sensed, inhibitory mode) implantation with the indication of complete heart block, and presented to us again 7 weeks later, with syncopal attacks. She admitted to repeatedly manipulating the pacemaker generator in her left pectoral region. Physical examination revealed a heart rate of 42 beats/minute, blood pressure 140/80 mmHg and bilateral crackles on lung auscultation. She had no cognitive deficit. An immediate electrocardiogram showed complete heart block with pacemaker spikes and failure to capture. Chest X-ray showed coiled and retracted right ventricular lead and rotated pulse generator. An emergent temporary pace maker was set at a rate of 60 beats per minute. Subsequently, she underwent successful lead repositioning with strong counselling to avoid further twiddling. Twiddler's syndrome should be considered as a cause of pacemaker failure in elderly patients presenting with bradyarrythmias following pacemaker implantation. Chest X-ray and electrocardiograms are simple and easily-available first line investigations for its diagnosis. Lead repositioning is required, however proper patient education and counselling against further manipulation is paramount to long-term management.

  14. Subacute right ventricle perforation by pacemaker lead presenting with left hemothorax and shock.

    Science.gov (United States)

    Nichols, Julianne; Berger, Natalie; Joseph, Praveen; Datta, Debapriya

    2015-01-01

    Cardiac perforation by pacemaker is a rare but potentially fatal complication. Acute perforations occurring within twenty-four hours of insertion of pacemaker can lead to hemopericardium, cardiac tamponade, and death. Hemothorax occurring as an acute complication of pacemaker insertion is reported but extremely rare. Previously, hemothorax and shock as a subacute complication following pacemaker insertion have not been reported. We report the case of an 85-year-old patient who presented with shock from hemothorax caused by pacemaker perforation, two weeks after insertion. Device interrogation showed normal function. Chest X-ray and echocardiogram missed lead dislocation and the diagnosis was made on computed tomogram (CT) of the chest. Following surgical repair, a new ventricular pacemaker was placed transvenously in the right ventricular septum. This case illustrates that CT scan of the chest should be performed in all patients in whom cardiac perforation by pacemaker is suspected but not diagnosed on chest X-ray and echocardiogram. Normal functioning of pacemaker on device interrogation does not exclude perforation.

  15. Subacute Right Ventricle Perforation by Pacemaker Lead Presenting with Left Hemothorax and Shock

    Directory of Open Access Journals (Sweden)

    Julianne Nichols

    2015-01-01

    Full Text Available Cardiac perforation by pacemaker is a rare but potentially fatal complication. Acute perforations occurring within twenty-four hours of insertion of pacemaker can lead to hemopericardium, cardiac tamponade, and death. Hemothorax occurring as an acute complication of pacemaker insertion is reported but extremely rare. Previously, hemothorax and shock as a subacute complication following pacemaker insertion have not been reported. We report the case of an 85-year-old patient who presented with shock from hemothorax caused by pacemaker perforation, two weeks after insertion. Device interrogation showed normal function. Chest X-ray and echocardiogram missed lead dislocation and the diagnosis was made on computed tomogram (CT of the chest. Following surgical repair, a new ventricular pacemaker was placed transvenously in the right ventricular septum. This case illustrates that CT scan of the chest should be performed in all patients in whom cardiac perforation by pacemaker is suspected but not diagnosed on chest X-ray and echocardiogram. Normal functioning of pacemaker on device interrogation does not exclude perforation.

  16. A Study of Clinicopathological Differences Between Right-sided and Left-sided Colon Cancers

    OpenAIRE

    芳賀, 駿介; 遠藤, 俊吾; 加藤, 博之; 高橋, 直樹; 吉松, 和彦; 橋本, 雅彦; 石橋, 敬一郎; 梅原, 有弘; 横溝, 肇; 梶原, 哲郎; Shunsuke, HAGA; Shungo, ENDO; Hiroyuki, KATO; Naoki, TAKAHASHI; Kazuhiko, YOSHIMATSU

    1996-01-01

    The present study was aimed to determine the clinicopathological features of cancers of the right-sided colon (cecum, ascending colon, transverse colon) and left-sided colon (descending colon, sigmoid colon) in order to help improve the efficacy of their treatment. Excluding multiple cancer cases, 364 patients with primary colon cancer underwent surgey at our department between 1974 and 1994; they comprised 171 individuals with right-sided colon cancer and 193 with left-sided colon cancer. A ...

  17. Left-Sided Patent Ductus Arteriosus in a Right-Sided Aortic Arch

    Directory of Open Access Journals (Sweden)

    Ming-Yen Ng

    2014-01-01

    Full Text Available We present a 31-year-old female with repaired tetralogy of Fallot (TOF and right-sided aortic arch (RAA with left-sided patent ductus arteriosus (PDA originating from the left brachiocephalic artery. This is a rare finding but most common site for a PDA in TOF and a RAA. To the best of our knowledge, this is the first demonstration of this rare finding on MRI in the literature.

  18. Left-sided gall bladder: Report of two cases

    Directory of Open Access Journals (Sweden)

    Chrungoo R

    2007-01-01

    Full Text Available Left-sided gall bladder without situs inversus viscerum is a rare albeit recognized clinical entity. We report our experience of two cases of left-sided gall bladder in two women aged 36 and 48 who underwent laparoscopic cholecystectomy for chronic calculous cholecystitis. Left-sided gall bladder may provide an unusual surprise to the surgeons during laparoscopy as routine pre-operative studies may not always detect the anomaly. Awareness of the unpredictable confluence of the cystic duct into the common bile duct (CBD and selective use of intraoperative cholangiography aid in the safe laparoscopic management of this unusual entity.

  19. Pacemaker therapy in low-birth-weight infants.

    Science.gov (United States)

    Fuchigami, Tai; Nishioka, Masahiko; Akashige, Toru; Shimabukuro, Atsuya; Nagata, Nobuhiro

    2018-02-01

    Infants born with complete atrioventricular block (CAVB) and fetal bradycardia are frequently born with low birth weight. Three low-birth-weight CAVB infants underwent temporary pacemaker implantation, followed by permanent single-chamber pacemaker implantation at median body weights of 1.7 and 3.2 kg, respectively. All infants caught up with their growth curves and had >3 years of estimated residual battery life. This two-stage strategy was successful in facilitating permanent pacemaker implantation in low-birth-weight babies. Placement of single-chamber pacemaker on the apex of the left ventricle appears to be associated with longer battery lifespan. © 2018 Wiley Periodicals, Inc.

  20. SU-F-T-654: Pacemaker Dose Estimate Using Optically Stimulated Luminescent Dosimeter for Left Breast Intraoperative Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Chen, Y; Goenka, A; Sharma, A; Wang, L; Cao, Y; Jamshidi, A [Northwell Health, Lake Success, NY (United States)

    2016-06-15

    Purpose: To assess and report the in vivo dose for a patient with a pacemaker being treated in left breast intraoperative radiation therapy (IORT). The ZEISS Intrabeam 50 kVp X-ray beam with a spherical applicator was used. Methods: The optically stimulated luminescent dosimeters (OSLDs) (Landauer nanoDots) were employed and calibrated under the conditions of the Intrabeam 50 kVp X-rays. The nanoDots were placed on the patient at approximately 15 cm away from the lumpectomy cavity both under and above a shield of lead equivalence 0.25 mm (RayShield X-Drape D-110) covering the pacemaker area during IORT with a 5 cm spherical applicator. Results: The skin surface dose near the pacemaker during the IORT with a prescription of 20 Gy was measured as 4.0±0.8 cGy. The dose behind the shield was 0.06±0.01 Gy, demonstrating more than 98% dose reduction. The in vivo skin surface doses during a typical breast IORT at a 4.5 cm spherical applicator surface were further measured at 5, 10, 15, and 20 cm away to be 159±11 cGy, 15±1 cGy, 6.6±0.5 cGy, and 1.8±0.1 cGy, respectively. A power law fit to the dose versus the distance z from the applicator surface yields the dose fall off at the skin surface following z^-2.5, which can be used to estimate skin doses in future cases. The comparison to an extrapolation of depth dose in water reveals an underestimate of far field dose using the manufactory provided data. Conclusion: The study suggests the appropriateness of OSLD as an in vivo skin dosimeter in IORT using the Intrabeam system in a wide dose range. The pacemaker dose measured during the left breast IORT was within a safe limit.

  1. Correlation analysis of the relationship between B-type natriuretic peptide and selected echocardiographic parameters in patients with permanent pacemakers

    Directory of Open Access Journals (Sweden)

    Janusz Sielski

    2016-01-01

    Full Text Available Introduction: The present study was undertaken to evaluate the practical value of BNP measurements and echocardiographic left ventricular volume index in patients with permanent pacemakers because there are no such reports in the literature. Aim of the research: The aim of the study was to reveal multiple correlations between BNP levels and selected echocardiographic parameters of the left atrium in patients with permanent pacemakers. In the literature there are reports on the significance of BNP values and left atrial size in patients with permanent pacemakers. The results of the present study appear to be of value in the outpatient assessment of these patients. Material and methods: We analysed a group of 117 patients with permanent pacemakers (AAI/R 21 patients, DDD/R 59 patients, VVI/R 37 patients and 48 healthy volunteers serving as the control group. BNP measurements were performed on venous blood samples using Triage meters. The Simpson method and the ellipse method were used to assess the left atrium on echocardiography. Results: There was a significant correlation between BNP and maximum left atrial volume, minimum left atrial volume, and left atrial volume index in patients with AAI/R, DDD/R, and VVI/R pacemakers at 3 and 6 months after the implantation. Conclusions : In patients after implantation of permanent pacemakers there are correlations between BNP values and echocardiographic left atrial parameters, especially in patients with DDD/R pacemakers. Left atrial function improves in patients with DDD/R pacemakers. Pacemaker check-up should be extended to include BNP measurements and echocardiographic assessment of the left atrium.

  2. Leadless Pacemakers: State of the Art and Future Perspectives.

    Science.gov (United States)

    Della Rocca, Domenico G; Gianni, Carola; Di Biase, Luigi; Natale, Andrea; Al-Ahmad, Amin

    2018-03-01

    Leadless pacemaker therapy is a new technology that aims at avoiding lead- and pocket-related complications of conventional transvenous and epicardial pacing. To date, 2 self-contained leadless pacemakers for right ventricular pacing have been clinically available: the Nanostim Leadless Pacemaker System and the Micra Transcatheter Pacing System. Additionally, a new multicomponent leadless pacemaker for endocardial left ventricular pacing has been proposed as an alternative choice for cardiac resynchronization therapy. In this review, we describe the state of the art of leadless pacing and compare the currently available devices with traditional transvenous leadless pacemakers. Copyright © 2017 Elsevier Inc. All rights reserved.

  3. Prevalence of left-sided melanomas in an Irish population.

    LENUS (Irish Health Repository)

    de Blacam, C

    2012-02-01

    BACKGROUND: A predominance of melanomas on the left side of the body has recently been described. No associations between tumour laterality and gender, age or anatomical site have been identified. AIM: The aim of this study was to investigate the prevalence of left-sided melanomas in an Irish population and to examine potential associations with various patient and tumour characteristics. METHODS: A retrospective chart review of patients with cutaneous melanoma who were treated over a 10-year period was carried out. Lateral distribution of melanoma on either side of the body was compared using chi(2) analysis and evaluated by gender, age group, anatomic location, histologic subtype and Breslow depth. RESULTS: More melanomas occurred on the left side (57%, P = 0.015), and this finding was particularly significant in females. For both genders combined, there were no statistically significant differences in laterality by age group, anatomic location, type of melanoma and Breslow depth. There were significantly more superficial spreading melanomas on the left side in both men and women. CONCLUSIONS: This study demonstrates a predominance of left-sided melanomas in Irish patients. While a number of demographic and molecular associations have been proposed, further research is required to fully explain this phenomenon.

  4. Prevalence of left-sided melanomas in an Irish population.

    LENUS (Irish Health Repository)

    de Blacam, C

    2011-04-17

    BACKGROUND: A predominance of melanomas on the left side of the body has recently been described. No associations between tumour laterality and gender, age or anatomical site have been identified. AIM: The aim of this study was to investigate the prevalence of left-sided melanomas in an Irish population and to examine potential associations with various patient and tumour characteristics. METHODS: A retrospective chart review of patients with cutaneous melanoma who were treated over a 10-year period was carried out. Lateral distribution of melanoma on either side of the body was compared using χ(2) analysis and evaluated by gender, age group, anatomic location, histologic subtype and Breslow depth. RESULTS: More melanomas occurred on the left side (57%, P = 0.015), and this finding was particularly significant in females. For both genders combined, there were no statistically significant differences in laterality by age group, anatomic location, type of melanoma and Breslow depth. There were significantly more superficial spreading melanomas on the left side in both men and women. CONCLUSIONS: This study demonstrates a predominance of left-sided melanomas in Irish patients. While a number of demographic and molecular associations have been proposed, further research is required to fully explain this phenomenon.

  5. Fallopian Tube Herniation through Left Sided Abdominal Drain Site.

    Science.gov (United States)

    Hussain, Khalid; Masood, Jovaria

    2016-06-01

    Intra-abdominal drains have been used since long to prevent intra-abdominal collection, and detect any anastomotic leaks. We report a case of left sided fallopian tube herniation from a left lower abdominal drain site in a 27-year female who underwent caesarian section for breach presentation. Several complications related to drain usage has been described but left sided fallopian tube prolapse through drain site has not been reported in literature.

  6. [Sport for pacemaker patients].

    Science.gov (United States)

    Israel, C W

    2012-06-01

    Sport activity is an important issue in many patients with a pacemaker either because they performed sport activities before pacemaker implantation to reduce the cardiovascular risk or to improve the course of an underlying cardiovascular disease (e.g. coronary artery disease, heart failure) by sports. Compared to patients with an implantable cardioverter defibrillator (ICD) the risks from underlying cardiovascular disease (e.g. ischemia, heart failure), arrhythmia, lead dysfunction or inappropriate therapy are less important or absent. Sport is contraindicated in dyspnea at rest, acute heart failure, new complex arrhythmia, acute myocarditis and acute myocardial infarction, valvular disease with indications for intervention and surgery and comorbidities which prevent physical activity. Patients with underlying cardiovascular disease (including hypertension) should preferably perform types and levels of physical activity that are aerobic (with dynamic exercise) such as running, swimming, cycling instead of sport with high anaerobic demands and high muscular workload. In heart failure, studies demonstrated advantages of isometric sport that increases the amount of muscle, thereby preventing cardiac cachexia. Sport with a risk of blows to the chest or physical contact (e.g. boxing, rugby, martial arts) should be avoided. Implantation, programming and follow-up should respect specific precautions to allow optimal physical activity with a pacemaker including implantation of bipolar leads on the side contralateral to the dominant hand, individual programming of the upper sensor and tracking rate and regular exercise testing.

  7. Left-sided portal hypertension revisited

    Directory of Open Access Journals (Sweden)

    Antonio Manenti

    2016-12-01

    Conclusions: In every case of left-sided portal hypertension, upper digestive endoscopy and close follow-up are recommended. Besides, computed tomography can demonstrate particular conditions directly favoring gastroesophageal varices, and aid in selection of the appropriate therapeutic decisions. [Arch Clin Exp Surg 2016; 5(4.000: 211-215

  8. Tricuspid valve repair for severe tricuspid regurgitation due to pacemaker leads.

    Science.gov (United States)

    Uehara, Kyokun; Minakata, Kenji; Watanabe, Kentaro; Sakaguchi, Hisashi; Yamazaki, Kazuhiro; Ikeda, Tadashi; Sakata, Ryuzo

    2016-07-01

    Tricuspid valve regurgitation due to pacemaker leads is a well-known complication. Although some reports have suggested that pacemaker leads should be surgically explanted, strongly adhered leads cannot always be removed. The aim of this study was to describe our tricuspid valve repair techniques with pacemaker leads left in situ. Our retrospective study investigated 6 consecutive patients who required tricuspid valve surgery for severe regurgitation induced by pacemaker leads. From the operative findings, we identified 3 patterns of tricuspid valve and pacemaker lead involvement. In 3 patients, the leads were caught in the chordae, in 2 patients, tricuspid regurgitation was caused by lead impingement on the septal leaflet, and in 3 patients, tricuspid valve leaflets had been perforated by the pacemaker leads. During surgery, all leads were left in situ after being separated from the leaflet or valvular apparatus. In addition, suture annuloplasty was performed for annular dilatation in all cases. In one patient, the lead was reaffixed to the annulus after the posterior leaflet was cut back towards the annulus, and the leaflet was then closed. There was one hospital death due to sepsis. The degree of tricuspid regurgitation was trivial in all surviving patients at discharge. During a mean follow-up of 21 months, one patient died from pneumonia 20 months after tricuspid valve repair. In patients undergoing tricuspid valve surgery due to severe tricuspid regurgitation caused by pacemaker leads, the leads can be left in situ after proper repair with annuloplasty. © The Author(s) 2016.

  9. Beware the left-sided gallbladder

    African Journals Online (AJOL)

    An abdominal ultrasound scan showed multiple stones in the gallbladder with no evidence of acute cholecystitis and a common bile duct (CBD) diameter of 4 mm. Owing to the atypical nature of her pain, a computed tomography (CT) scan was requested which showed a left-sided gallbladder containing gallstones (Fig. 1).

  10. Placement of the Left Side AED Pad is Poor: Training on the Left Compared to the Right Side of a Manikin Does Not Improve Pad Placement

    DEFF Research Database (Denmark)

    Stærk, Mathilde; Bødtker, Henrik; Rahbek, Søren

    2015-01-01

    participating in a first aid course were randomized to learn automated external defibrillation sitting on the left or right side of a manikin during AED training. After course completion participants operated a training AED (Lifepak® CR-T AED Trainer, PhysioControl) and placed AED pads according to instructions...... volunteers were included and randomized to AED training on the left (n=14, 43% male, age: 47.9 years) and right (n=16, 25% male, age: 46.7 years) side of a manikin. There was no difference in left pad placement when trained on the left or right side (distance to recommended left apical pad position (mean...

  11. Comparison between the incidence of right and left sided congenital torticollis

    International Nuclear Information System (INIS)

    Aslam, S.; Bashir, M.S.; Hussain, S.I.

    2013-01-01

    Congenital torticollis is an intriguing condition of unknown origin, characterized by unilateral shortening and tightness of the sternocleidomastoid muscle. Patients usually present with head tilt, facial asymmetry and plagiocephaly. A sternomastoid mass or tumor may or may not be clinically apparent. Untreated, cervical function and facial cosmesis may be severely compromised. Objective: My study is aimed at establishing a comparison between the incidence of right versus left sided congenital torticollis. Method: This observational study included 30 patients of congenital torticollis that completed the questionnaire. The data was collected from patients coming to the Physiotherapy and Orthopedic departments of Children Hospital, Lahore. Results: Results showed that right side was involved in 19 (63.3%) patients and left side was involved in 11 (36.7%) patients. Out of 30 patients, 14 (46.7%) were male, of which 8 had right sided congenital torticollis and 6 had left sided congenital torticollis, and 16 (53.3%) were female, of which right sided congenital torticollis and 5 had left sided congenital torticollis. Conclusion: Hence it is concluded that incidence of right sided congenital torticollis is more common than left sided congenital torticollis. The incidence of con-genital torticollis is higher in females than in males. (author)

  12. Mapping of the left-sided phrenic nerve course in patients undergoing left atrial catheter ablations.

    Science.gov (United States)

    Huemer, Martin; Wutzler, Alexander; Parwani, Abdul S; Attanasio, Philipp; Haverkamp, Wilhelm; Boldt, Leif-Hendrik

    2014-09-01

    Catheter ablation of atrial fibrillation has been associated with left-sided phrenic nerve palsy. Knowledge of the individual left phrenic nerve course therefore is essential to prevent nerve injury. The aim of this study was to test the feasibility of an intraprocedural pace mapping and reconstruction of the left phrenic nerve course and to characterize which anatomical areas are affected. In patients undergoing left atrial catheter ablation, a three-dimensional map of the left atrial anatomical structures was created. The left-sided phrenic nerve course was determined by high-output pace mapping and reconstructed in the map. In this study, 40 patients with atrial fibrillation or atrial tachycardias were included. Left phrenic nerve capture was observed in 23 (57.5%) patients. Phrenic nerve was captured in 22 (55%) patients inside the left atrial appendage, in 22 (55%) in distal parts, in 21 (53%) in medial parts, and in two (5%) in ostial parts of the appendage. In three (7.5%) patients, capture was found in the distal coronary sinus and in one (2.5%) patient in the left atrium near the left atrial appendage ostium. Ablation target was changed due to direct spatial relationship to the phrenic nerve in three (7.5%) patients. No phrenic nerve palsy was observed. Left-sided phrenic nerve capture was found inside and around the left atrial appendage in the majority of patients and additionally in the distal coronary sinus. Phrenic nerve mapping and reconstruction can easily be performed and should be considered prior catheter ablations in potential affected areas. ©2014 Wiley Periodicals, Inc.

  13. Heart pacemaker

    Science.gov (United States)

    Cardiac pacemaker implantation; Artificial pacemaker; Permanent pacemaker; Internal pacemaker; Cardiac resynchronization therapy; CRT; Biventricular pacemaker; Arrhythmia - pacemaker; Abnormal heart ...

  14. Troponin T elevation after permanent pacemaker implantation.

    Science.gov (United States)

    Chen, Xueying; Yu, Ziqing; Bai, Jin; Hu, Shulan; Wang, Wei; Qin, Shengmei; Wang, Jingfeng; Sun, Zhe; Su, Yangang; Ge, Junbo

    2017-08-01

    The objective of the study is to study the incidence, significance, and factors associated with cardiac troponin T (CTNT) elevation after pacemaker implantation. Three hundred seventy-four patients (104 single-chamber pacemakers or ICD, 243 dual-chamber pacemakers, and 27 cardiac resynchronization therapy/cardiac resynchronization therapy defibrillator) who had normal levels of CTNT at baseline and underwent implantation of a permanent pacemaker system were included in this study. Serum levels of CTNT were measured at baseline, 6 and 24 h after the implantation procedure. The median of CTNT levels increased from 0.012 ng/mL at baseline to 0.032 and 0.019 ng/mL at 6 and 24 h after the procedure, respectively (all p 0.09 ng/mL). After 1-year follow-up, the incidence of complications including dislodgement of the lead, pocket infection, pneumothorax, hemothorax, and vein thrombus and cardiac outcomes including hospitalization of heart failure, coronary artery disease, arrhythmia, and cardiovascular mortality was not significantly different between the normal and elevated CTNT groups at 6 h after the procedure. By logistic regression analysis, gender, N-terminal pro-B type natriuretic peptide (NT-pro-BNP) at baseline, left ventricular ejection fractions (LVEF), estimated glomerular filtration rate (eGFR), and fluoroscopy time were independently associated with CTNT elevation after adjusted for age, pacemaker types, right ventricle lead location (RVA or RVOT), heart function, and left ventricular end systolic dimension. Pacemaker implantation was found to be accompanied with CTNT elevation in 55.6% of the patients at 6 h after the procedure, and its kinetics were fast, which might not be related to the complications and adverse cardiac outcomes within 1 year of follow-up. Moreover, gender, NT-pro-BNP at baseline, LVEF, eGFR, and fluoroscopy time were found to be independent predictors of CTNT elevation.

  15. Left Sided Amyand's Hernia, A Rare Occurance: A Case Report.

    Science.gov (United States)

    Ravishankaran, Praveen; Mohan, G; Srinivasan, A; Ravindran, G; Ramalingam, A

    2013-06-01

    This is a case report about a 35 year old man admitted with complains of obstructed left sided inguinal hernia. On exploration of the left inguinal canal to our surprise a normal appendix was found in addition to a gangrenous omentum. Resection of the gangrenous omentum was done. Appendectomy was done. This case is reported for its rare occurance as only three such cases of left sided amyand's hernia has been reported so far in literature[4-6].

  16. Quality of life differences in patients with right- versus left-sided facial paralysis: Universal preference of right-sided human face recognition.

    Science.gov (United States)

    Ryu, Nam Gyu; Lim, Byung Woo; Cho, Jae Keun; Kim, Jin

    2016-09-01

    We investigated whether experiencing right- or left-sided facial paralysis would affect an individual's ability to recognize one side of the human face using hybrid hemi-facial photos by preliminary study. Further investigation looked at the relationship between facial recognition ability, stress, and quality of life. To investigate predominance of one side of the human face for face recognition, 100 normal participants (right-handed: n = 97, left-handed: n = 3, right brain dominance: n = 56, left brain dominance: n = 44) answered a questionnaire that included hybrid hemi-facial photos developed to determine decide superiority of one side for human face recognition. To determine differences of stress level and quality of life between individuals experiencing right- and left-sided facial paralysis, 100 patients (right side:50, left side:50, not including traumatic facial nerve paralysis) answered a questionnaire about facial disability index test and quality of life (SF-36 Korean version). Regardless of handedness or hemispheric dominance, the proportion of predominance of the right side in human face recognition was larger than the left side (71% versus 12%, neutral: 17%). Facial distress index of the patients with right-sided facial paralysis was lower than that of left-sided patients (68.8 ± 9.42 versus 76.4 ± 8.28), and the SF-36 scores of right-sided patients were lower than left-sided patients (119.07 ± 15.24 versus 123.25 ± 16.48, total score: 166). Universal preference for the right side in human face recognition showed worse psychological mood and social interaction in patients with right-side facial paralysis than left-sided paralysis. This information is helpful to clinicians in that psychological and social factors should be considered when treating patients with facial-paralysis. Copyright © 2016 British Association of Plastic, Reconstructive and Aesthetic Surgeons. Published by Elsevier Ltd. All rights reserved.

  17. Isolated left-sided pulmonary artery agenesis with left lung hypoplasia: A report of two cases

    Directory of Open Access Journals (Sweden)

    V Govindaraj

    2017-01-01

    Full Text Available Unilateral absence of pulmonary artery or pulmonary artery agenesis (UAPA is a rare congenital malformation that can present as an isolated lesion or in association with other cardiac anomalies. Though congenital, presentation in adults are also reported. Most common presentation in adults is of exercise intolerance. The developing lung on the affected side is hypoplastic. Diagnosis of UAPA is established by imaging methods like CT and MRI . There is no specific treatment for this condition. Treatment depends on patients symptomatology, presence of pulmonary hypertension and collateral circulation. Presence of pulmonary hypertension carries a bad prognosis. We present two adult patients with isolated left sided unilateral pulmonary artery agenesis with ipsilateral lung hypoplasia. The diagnosis was confirmed by CT chest and perfusion scan.

  18. Discuss the cause and treatment of pacemaker lead dislocation and deal with

    International Nuclear Information System (INIS)

    Chen Yueguang; Zhang Dadong; Lu Jie; Yang Hui; Liu Chunyan; Zhang Wei

    2003-01-01

    Objective: To follow up the patients with pacemaker, observe the condition of pacemaker lead, to explore the cause of lead dislocation, to find out and prevent its occurrence. Methods: Summarizing the clinical data of 6 patients with pacemaker, 7 pacemaker leads with 8 time dislocation, pacemaker 2 DDDR, 2 DDD, 2 VVI. Results: Four patients were punctured from right subclavian vein, one from left subclavian vein and one from right brachiocephalic vein; four leads were dislocation in atrium and one mildly dislocation; four leads dislocation in ventricle and two mildly dislocation; There were 3 old women with 4 leads and 5 times of dislocation

  19. Left-side changes in thoracic X-ray follow-ups after cardiosurgery

    Energy Technology Data Exchange (ETDEWEB)

    Witte, G.; Buecheler, E.; Grabbe, E.; Darup, J.

    1983-03-01

    In routine thoracic X-ray follow-ups after cardiosurgical procedures, 93.2% of our patients (n=88) showed radiologically perceivable pathological changes, reduced ventilation being the most common phenomenon. 77.1% of the patients showed signs of reduced ventilation, mostly the left side. Limited motility of the diaphragm on the left side was visible in 69% of the cases studied. The defective motility of the left diaphragm is attributable to direct damage to the left nervus phrenicus caused by extracardial heart cooling during surgery.

  20. Left-side changes in thoracic X-ray follow-ups after cardiosurgery

    International Nuclear Information System (INIS)

    Witte, G.; Buecheler, E.; Grabbe, E.; Darup, J.

    1983-01-01

    In routine thoracic X-ray follow-ups after cardiosurgical procedures, 93.2% of our patients (n=88) showed radiologically perceivable pathological changes, reduced ventilation being the most common phenomenon. 77.1% of the patients showed signs of reduced ventilation, mostly the left side. Limited motility of the diaphragm on the left side was visible in 69% of the cases studied. The defective motility of the left diaphragm is attributable to direct damage to the left nervus phrenicus caused by extracardial heart cooling during surgery. (orig.) [de

  1. Neglecting the Left Side of a City Square but Not the Left Side of Its Clock: Prevalence and Characteristics of Representational Neglect

    Science.gov (United States)

    Guariglia, Cecilia; Palermo, Liana; Piccardi, Laura; Iaria, Giuseppe; Incoccia, Chiara

    2013-01-01

    Representational neglect, which is characterized by the failure to report left-sided details of a mental image from memory, can occur after a right hemisphere lesion. In this study, we set out to verify the hypothesis that two distinct forms of representational neglect exist, one involving object representation and the other environmental representation. As representational neglect is considered rare, we also evaluated the prevalence and frequency of its association with perceptual neglect. We submitted a group of 96 unselected, consecutive, chronic, right brain-damaged patients to an extensive neuropsychological evaluation that included two representational neglect tests: the Familiar Square Description Test and the O'Clock Test. Representational neglect, as well as perceptual neglect, was present in about one-third of the sample. Most patients neglected the left side of imagined familiar squares but not the left side of imagined clocks. The present data show that representational neglect is not a rare disorder and also support the hypothesis that two different types of mental representations (i.e. topological and non-topological images) may be selectively damaged in representational neglect. PMID:23874416

  2. Mean size estimation yields left-side bias: Role of attention on perceptual averaging.

    Science.gov (United States)

    Li, Kuei-An; Yeh, Su-Ling

    2017-11-01

    The human visual system can estimate mean size of a set of items effectively; however, little is known about whether information on each visual field contributes equally to the mean size estimation. In this study, we examined whether a left-side bias (LSB)-perceptual judgment tends to depend more heavily on left visual field's inputs-affects mean size estimation. Participants were instructed to estimate the mean size of 16 spots. In half of the trials, the mean size of the spots on the left side was larger than that on the right side (the left-larger condition) and vice versa (the right-larger condition). Our results illustrated an LSB: A larger estimated mean size was found in the left-larger condition than in the right-larger condition (Experiment 1), and the LSB vanished when participants' attention was effectively cued to the right side (Experiment 2b). Furthermore, the magnitude of LSB increased with stimulus-onset asynchrony (SOA), when spots on the left side were presented earlier than the right side. In contrast, the LSB vanished and then induced a reversed effect with SOA when spots on the right side were presented earlier (Experiment 3). This study offers the first piece of evidence suggesting that LSB does have a significant influence on mean size estimation of a group of items, which is induced by a leftward attentional bias that enhances the prior entry effect on the left side.

  3. Left-sided native valve Staphylococcus aureus endocarditis

    NARCIS (Netherlands)

    Slabbekoorn, M.; Horlings, H. M.; van der Meer, J. T. M.; Windhausen, A.; Van der Sloot, J. A. P.; Lagrand, W. K.

    2010-01-01

    Despite improved diagnostic tools and expanded treatment options, left-sided native valve endocarditis caused by Staphylococcus aureus infection remains a serious and destructive disease. The high morbidity and mortality, however, can be reduced by early recognition, correct diagnosis, and

  4. Midgut malrotation presenting with left-sided acute appendicitis and CT inversion sign

    Science.gov (United States)

    Çağlar, Emrah; Arıbaş, Bilgin; Tiken, Ramazan; Keskin, Suat

    2014-01-01

    In patients presenting with abdominal pain, appendicitis is the most common surgical disorder. Appendicitis causing pain in the left lower quadrant is extremely rare and can occur with congenital abnormalities that include true left-sided appendix or as an atypical presentation of right-sided long appendix, which projects into the left lower quadrant. We report a case of a 69-year-old man showing midgut malrotation with acute appendicitis presenting as left lower quadrant abdominal pain. PMID:24682135

  5. Isolated left-sided partial anomalous pulmonary venous connection in a child.

    Science.gov (United States)

    Onan, İsmihan Selen; Sen, Onur; Gökalp, Selman; Onan, Burak

    2017-09-01

    Isolated left-sided partial anomalous pulmonary venous connection with intact interatrial septum is a rare diagnosis in childhood. In these cases, a vertical vein drains the left upper pulmonary lobe into the brachiocephalic vein and finally to the right atrium. Surgical treatment is performed to prevent right ventricular failure and pulmonary artery disease in advanced age. In this report, the rare entity of isolated left-sided anomalous pulmonary venous connection in a 14-year-old girl and successful minimally invasive surgery without cardiopulmonary bypass are described.

  6. A disappearing left-sided neck mass.

    Science.gov (United States)

    Lutwak, Nancy; Dill, Curt

    2012-07-04

    The patient was a 48-year-old man who presented to the emergency department with complaints of a left-sided painful neck mass, which changed in size relative to ingestion of meals. He denied voice change, fever, chills, weight loss, dysphagia and hoarseness. Physical examination was unremarkable. CT scan demonstrated a 3.9 mm calculus of the submandibular gland duct. Therapeutic sialendocopy was successfully performed.

  7. Randomized placebo controlled blinded study to assess valsartan efficacy in preventing left ventricle remodeling in patients with dual chamber pacemaker--Rationale and design of the trial.

    Science.gov (United States)

    Tomasik, Andrzej; Jacheć, Wojciech; Wojciechowska, Celina; Kawecki, Damian; Białkowska, Beata; Romuk, Ewa; Gabrysiak, Artur; Birkner, Ewa; Kalarus, Zbigniew; Nowalany-Kozielska, Ewa

    2015-05-01

    Dual chamber pacing is known to have detrimental effect on cardiac performance and heart failure occurring eventually is associated with increased mortality. Experimental studies of pacing in dogs have shown contractile dyssynchrony leading to diffuse alterations in extracellular matrix. In parallel, studies on experimental ischemia/reperfusion injury have shown efficacy of valsartan to inhibit activity of matrix metalloproteinase-9, to increase the activity of tissue inhibitor of matrix metalloproteinase-3 and preserve global contractility and left ventricle ejection fraction. To present rationale and design of randomized blinded trial aimed to assess whether 12 month long administration of valsartan will prevent left ventricle remodeling in patients with preserved left ventricle ejection fraction (LVEF ≥ 40%) and first implantation of dual chamber pacemaker. A total of 100 eligible patients will be randomized into three parallel arms: placebo, valsartan 80 mg/daily and valsartan 160 mg/daily added to previously used drugs. The primary endpoint will be assessment of valsartan efficacy to prevent left ventricle remodeling during 12 month follow-up. We assess patients' functional capacity, blood plasma activity of matrix metalloproteinases and their tissue inhibitors, NT-proBNP, tumor necrosis factor alpha, and Troponin T. Left ventricle function and remodeling is assessed echocardiographically: M-mode, B-mode, tissue Doppler imaging. If valsartan proves effective, it will be an attractive measure to improve long term prognosis in aging population and increasing number of pacemaker recipients. ClinicalTrials.org (NCT01805804). Copyright © 2015 Elsevier Inc. All rights reserved.

  8. Impact of patient-specific factors, irradiated left ventricular volume, and treatment set-up errors on the development of myocardial perfusion defects after radiation therapy for left-sided breast cancer

    International Nuclear Information System (INIS)

    Evans, Elizabeth S.; Prosnitz, Robert G.; Yu Xiaoli; Zhou Sumin; Hollis, Donna R.; Wong, Terence Z.; Light, Kim L.; Hardenbergh, Patricia H.; Blazing, Michael A.; Marks, Lawrence B.

    2006-01-01

    Purpose: The aim of this study was to assess the impact of patient-specific factors, left ventricle (LV) volume, and treatment set-up errors on the rate of perfusion defects 6 to 60 months post-radiation therapy (RT) in patients receiving tangential RT for left-sided breast cancer. Methods and Materials: Between 1998 and 2005, a total of 153 patients were enrolled onto an institutional review board-approved prospective study and had pre- and serial post-RT (6-60 months) cardiac perfusion scans to assess for perfusion defects. Of the patients, 108 had normal pre-RT perfusion scans and available follow-up data. The impact of patient-specific factors on the rate of perfusion defects was assessed at various time points using univariate and multivariate analysis. The impact of set-up errors on the rate of perfusion defects was also analyzed using a one-tailed Fisher's Exact test. Results: Consistent with our prior results, the volume of LV in the RT field was the most significant predictor of perfusion defects on both univariate (p = 0.0005 to 0.0058) and multivariate analysis (p = 0.0026 to 0.0029). Body mass index (BMI) was the only significant patient-specific factor on both univariate (p = 0.0005 to 0.022) and multivariate analysis (p = 0.0091 to 0.05). In patients with very small volumes of LV in the planned RT fields, the rate of perfusion defects was significantly higher when the fields set-up 'too deep' (83% vs. 30%, p = 0.059). The frequency of deep set-up errors was significantly higher among patients with BMI ≥25 kg/m 2 compared with patients of normal weight (47% vs. 28%, p = 0.068). Conclusions: BMI ≥25 kg/m 2 may be a significant risk factor for cardiac toxicity after RT for left-sided breast cancer, possibly because of more frequent deep set-up errors resulting in the inclusion of additional heart in the RT fields. Further study is necessary to better understand the impact of patient-specific factors and set-up errors on the development of RT

  9. EDUCATIONAL PECULIARITIES AND DIFFICULTIES OF CHILDREN WITH LEFT-SIDED LATERALITY: THE TECHNOLOGICAL SOLUTION OF THE PROBLEM

    Directory of Open Access Journals (Sweden)

    Maria Sitnikova

    2011-09-01

    Full Text Available Nowadays there is a significant increase of the incidence of left-handedness and sinistrality among schoolchildren. Theydemonstrate a large number of left-sided motor and sensory preferences which are considered as external markers offunctional hemispheric asymmetry of the brain. The purposes of this study are to investigate gender peculiarities and specificityof age-related dynamics of laterality pattern’s formation in junior schoolchildren and to find out educational peculiarities anddifficulties of left-handed children. The findings show that left-handers differ greatly in their mental development by havingsome peculiarities of intelligence, world’s perception and prevailing thinking strategies, ways of memorization, specificity ofemotional-affective expression. The main problems of left-handed children in school performance are academic failure, lack ofperseverance, anxiety neurosis, and extreme emotional lability. Integrated development of the left hemisphere and the righthemisphere thinking of left-handed schoolchildren is a favorable condition for harmonious personal and intellectualdevelopment and effective mastering of various modules of the school curriculum. The technological solution of the problem ofteaching the children with left-sided laterality is to include in educational programs some special exercises to developimagination, emotional sensitivity, integrity of perception, global view to the problems, creativeness, and original approachesto tasks’ solving. So a complex program for the intensive development of the right hemisphere of children who demonstrateleft-sided laterality to overcome the possible failure at primary school is proposed in this paper.

  10. Heart pacemaker - discharge

    Science.gov (United States)

    Cardiac pacemaker implantation - discharge; Artificial pacemaker - discharge; Permanent pacemaker - discharge; Internal pacemaker - discharge; Cardiac resynchronization therapy - discharge; CRT - discharge; ...

  11. Final generic environmental statement on the routine use of plutonium-powered cardiac pacemakers. Update of information on power sources for pacemakers

    International Nuclear Information System (INIS)

    1979-05-01

    The Final Environmental Statement on Routine Use of Plutonium-Powered Cardiac Pacemakers (FES) was issued in July 1976. Supplement 1, prepared in 1978, updates the FES with respect to power sources for pacemakers. Particular attention is given to the non-nuclear lithium batteries as alternatives to 238-Pu power sources in pacemakers. Supplement 1 also considers the current extent of pacemaker use and makeup of the patient population and concludes that the FES' conclusion is still valid that distribution of 238-Pu powered pacemakers for routine use should be authorized subject to specific conditions

  12. Pacemaker diagnostics in atrial fibrillation: limited usefulness for therapy initiation in a pacemaker practice.

    Science.gov (United States)

    Yedlapati, Neeraja; Fisher, John D

    2014-09-01

    We aimed to determine the practical value of pacemaker diagnostics for atrial fibrillation (AF) in an unselected general pacemaker practice, specifically workflow and initiation of anticoagulation or antiarrhythmic drug (AAD) therapy. We prospectively followed consecutive pacemaker interrogations over a period of 1 year to identify patients with AF (burden from 1% to 99%). We contacted referring physicians with AF details, and then determined whether the information resulted in therapeutic changes. Of the 1,100 pacemakers interrogated, 728 were dual chamber (DDDs) with AF diagnostic capability. AF was recorded in 73 (10%) but seven had limited information, leaving 66 patients; of these, 42 (63%) patients were already anticoagulated and in five (7%) patients, anticoagulation had been stopped because of complications. Initial diagnosis of AF was made by the pacemaker in 17 patients (26% of 66; 2% of 728); four (6% of 66) patients were newly initiated on anticoagulation. Of the 66 patients, 17 patients were already on AADs; 49 (74%) had satisfactory rate control or had other issues; only two (3% of 66; 0.3% of 728) received new AADs. Of 728 patients with DDD pacemakers, only 17 were newly discovered to have AF, and six (0.8%) had changes in medications based on the pacemaker data. Adding pacemaker-derived data to existing clinical information had little therapeutic impact, due to a combination of cumbersome workflow, and because AF was usually known to practitioners. Developments in automated monitoring systems may provide more accessible and therapeutically useful information. ©2014 Wiley Periodicals, Inc.

  13. Surgical management of colonic diverticular disease: discrepancy between right- and left-sided diseases.

    Science.gov (United States)

    Oh, Heung-Kwon; Han, Eon Chul; Ha, Heon-Kyun; Choe, Eun Kyung; Moon, Sang Hui; Ryoo, Seung-Bum; Jeong, Seung-Yong; Park, Kyu Joo

    2014-08-07

    To compare the outcome of the surgical management of left-sided and right-sided diverticular disease. The medical records of 77 patients who were surgically treated for diverticular disease between 1999 and 2010 in a tertiary referral hospital were retrospectively reviewed. The study population was limited to cases wherein the surgical specimen was confirmed as diverticulosis by pathology. Right-sided diverticula were classified as those arising from the cecum, ascending colon, and transverse colon, and those from the descending colon, sigmoid colon, and rectum were classified as left-sided diverticulosis. To assess the changing trend of occurrence of diverticulosis, data were compared with two previous studies of 51 patients. The proportion of left-sided disease cases was significantly increased compared to the results of our previous studies in 1994 and 2001, (27.5% vs 48.1%, P disease. However, patients with right-sided disease were significantly younger (50.9 year vs 64.0 year, P disease was significantly associated with a higher incidence of complicated diverticulitis (89.2% vs 57.5%, P diverticular disease, the incidence of left-sided disease in Korea has increased since 2001 and is associated with worse surgical outcomes.

  14. Combined atresia of one left-sided and one right-sided cardiac valve in a premature newborn.

    Science.gov (United States)

    Roberts, William C; Sing, Alan C; Guileyardo, Joseph M

    2017-10-01

    Described herein is the heart of a 2-day-old newborn, the product of a 25-week gestation, with atresia of two cardiac valves, one on the right side and one on the left side, apparently a previously undescribed entity.

  15. Sleep on the right side-Get cancer on the left?

    Science.gov (United States)

    Hallberg, Orjan; Johansson, Olle

    2010-06-01

    Breast cancer frequently occurs in the left breast among both women and men [R. Roychoudhuri, V. Putcha, H. Møller, Cancer and laterality: a study of the five major paired organs (UK), Cancer Causes Control 17 (2006) 655-662; M.T. Goodman, K.H. Tung, L.R. Wilkens, Comparative epidemiology of breast cancer among men and women in the US, 1996 to 2000, Cancer Causes Control 17 (2006) 127-136; C.I. Perkins, J. Hotes, B.A. Kohler, H.L. Howe, Association between breast cancer laterality and tumor location, United States, 1994-1998, Cancer Causes Control 15 (2004) 637-645; H.A. Weiss, S.S. Devesa, L.A. Brinton, Laterality of breast cancer in the United States, Cancer Causes Control 7 (1996) 539-543; A. Ekbom, H.O. Adami, D. Trichopoulos, M. Lambe, C.C. Hsieh, J. Pontén, Epidemiologic correlates of breast cancer laterality (Sweden), Cancer Causes Control 5 (1994) 510-516]. Moreover, recent results showed that the left side of the body is more prone to melanoma than the right side [D.H. Brewster, M.J. Horner, S. Rowan, P. Jelfs, E. de Vries, E. Pukkala, Left-sided excess of invasive cutaneous melanoma in six countries, Eur. J. Cancer 43 (2007) 2634-2637]. Current explanations for left-sided breast cancer include handedness [L. Titus-Ernstoff, P.A. Newcomb, K.M. Egan, et al., Left-handedness in relation to breast cancer risk in postmenopausal women, Epidemiology 11 (2000) 181-184; M.A. Kramer, S. Albrecht, R.A. Miller, Handedness and the laterality of breast cancer in women, Nurs. Res. 34 (1985) 333-337; M.K. Ramadhani, S.G. Elias, P.A. van Noord, D.E. Grobbee, P.H. Peeters, C.S. Uiterwaal, Innate left handedness and risk of breast cancer: case-cohort study, BMJ 331 (2005) 882-883], size difference, nursing preference, and brain structure. However, men are affected even more by left laterality than women, thus many of these explanations are unconvincing. Increasing rates of skin melanoma have been associated with immune-disruptive radiation from FM/TV transmitters [O

  16. Radiation therapy planning of a breast cancer patient with in situ pacemaker-challenges and lessons

    Energy Technology Data Exchange (ETDEWEB)

    Munshi, Anusheel; Wadasadawala, Tabassum; Budrukkar, Ashwini; Jalali, Rakesh; Dinshaw, Ketayun A [Dept. of Radiation Oncology, Tata Memorial Hospital, Mumbai (India); Sharma, Pramod Kumar; Sharma, Dayananda [Dept. of Radiation Physics, Tata Memorial Hospital, Mumbai (India)

    2008-02-15

    A postmenopausal lady with an in situ pacemaker developed a lump in the left breast and was diagnosed to have breast cancer. The patient underwent breast conservative surgery and was planned for post operative radiotherapy. The location of the tumor relative to the pacemaker provided a unique challenge in planning radiotherapy and the patient had an uneventful post radiotherapy course. A literature review revealed that modern generation pacemakers are very sensitive to radiation compared to their older counterparts. The present article makes suggestions towards reducing dose in radiotherapy planning in pacemaker patients

  17. Radiation therapy planning of a breast cancer patient with in situ pacemaker-challenges and lessons

    International Nuclear Information System (INIS)

    Munshi, Anusheel; Wadasadawala, Tabassum; Budrukkar, Ashwini; Jalali, Rakesh; Dinshaw, Ketayun A.; Sharma, Pramod Kumar; Sharma, Dayananda

    2008-01-01

    A postmenopausal lady with an in situ pacemaker developed a lump in the left breast and was diagnosed to have breast cancer. The patient underwent breast conservative surgery and was planned for post operative radiotherapy. The location of the tumor relative to the pacemaker provided a unique challenge in planning radiotherapy and the patient had an uneventful post radiotherapy course. A literature review revealed that modern generation pacemakers are very sensitive to radiation compared to their older counterparts. The present article makes suggestions towards reducing dose in radiotherapy planning in pacemaker patients

  18. Perceptual expertise: can sensorimotor experience change holistic processing and left-side bias?

    Science.gov (United States)

    Tso, Ricky Van-yip; Au, Terry Kit-fong; Hsiao, Janet Hui-wen

    2014-09-01

    Holistic processing and left-side bias are both behavioral markers of expert face recognition. By contrast, expert recognition of characters in Chinese orthography involves left-side bias but reduced holistic processing, although faces and Chinese characters share many visual properties. Here, we examined whether this reduction in holistic processing of Chinese characters can be better explained by writing experience than by reading experience. Compared with Chinese nonreaders, Chinese readers who had limited writing experience showed increased holistic processing, whereas Chinese readers who could write characters fluently showed reduced holistic processing. This result suggests that writing and sensorimotor experience can modulate holistic-processing effects and that the reduced holistic processing observed in expert Chinese readers may depend mostly on writing experience. However, both expert writers and writers with limited experience showed similarly stronger left-side bias than novices did in processing mirror-symmetric Chinese characters; left-side bias may therefore be a robust expertise marker for object recognition that is uninfluenced by sensorimotor experience. © The Author(s) 2014.

  19. Psychosocial responses of children to cardiac pacemakers.

    Science.gov (United States)

    Alpern, D; Uzark, K; Dick, M

    1989-03-01

    To examine the psychosocial responses of children and adolescents with a cardiac pacemaker and compare their responses to those of their peers, we evaluated 30 pediatric pacemaker patients, aged 7 to 19 years, and two age- and sex-matched comparison groups, including 30 patients with similar heart disease but without pacemakers and 30 physically healthy children, using standardized psychometric tests and a specific interview format. We postulated that children with pacemakers would experience greater stress in psychosocial adaptation. No significant differences on standardized measures of trait anxiety, self-competence, or self-esteem were found between the pacemaker group and the comparison groups. In contrast, pacemaker subjects were significantly (p less than 0.05) more external in their locus-of-control orientation than were healthy subjects, suggesting a diminished sense of personal control and less autonomy. Pacemaker subjects, particularly the older ones, had significantly (p less than 0.05) greater knowledge of pacemaker systems than did subjects in the other two groups, facilitating the use of intellectualization as a coping mechanism. The pacemaker patients were likely to be as fearful of social rejection as of potential pacemaker failure. All three groups identified potential negative peer reactions toward an individual with a pacemaker. The patients with cardiac disease but without pacemakers and the healthy subjects perceived significant (p less than 0.05) social and emotional differences between patients with pacemakers and their peers, but the pacemaker patients did not view themselves as different from their peers. This study demonstrates healthy psychosocial adaptation of children with cardiac pacemakers. Although these children appear to cope effectively with the stress of their life situation through the use of denial and intellectualization, they may experience problems both in the development of autonomy and in social isolation and rejection.

  20. Differences between right- and left-sided colon cancer in patient characteristics, cancer morphology and histology.

    Science.gov (United States)

    Nawa, Toru; Kato, Jun; Kawamoto, Hirofumi; Okada, Hiroyuki; Yamamoto, Hiroshi; Kohno, Hiroyuki; Endo, Hisayuki; Shiratori, Yasushi

    2008-03-01

    Recently, the clinical and biological differences between right- and left-sided colon cancers have been widely debated. However, close analyses of these clinical differences, based on large-scale studies, have been scarcely reported. A total of 3552 consecutive Japanese colorectal cancer cases were examined and the clinical differences between right- and left-sided colon cancer cases were investigated. The proportion of right-sided colon cancer was relatively high in patients aged less than 40 years (33%) and more than 80 years (43%). The proportion of right-sided colon cancer in patients aged 40-59 years was relatively low (male 22% and female 29%). In male patients the proportion increased in the 70-79 years age group (30%), while in female patients the proportion increased in the 60-69 years age group (39%). Right-sided colon cancer was more likely to be detected at an advanced stage (T1 stage; left 22%, right 15%) (P cancer was dominant in the left colon (left 59%; right 40%) (P cancer in the right colon was significantly higher than that in the left colon (left 25%; right 44%) (P colon cancer was observed and the difference between male and female patients was highlighted. Other clinical features also differed between right- and left-sided colon cancer, suggesting that different mechanisms may be at work during right and left colon carcinogenesis.

  1. Conduction disturbances after TAVR: Electrophysiological studies and pacemaker dependency.

    Science.gov (United States)

    Makki, Nader; Dollery, Jenn; Jones, Danielle; Crestanello, Juan; Lilly, Scott

    Permanent pacemaker (PPM) placement occurs in 5-20% of patients after transcatheter aortic valve replacement (TAVR). Although predictors of pacemaker implantation have been established, features that predispose patients to pacemaker utilization on follow up have not been widely reported. We performed a retrospective review of patients undergoing commercial TAVR between 2011 and 2016. We collated patients that underwent in-hospital PPM implantation and had a follow up of at least 3months. Data abstraction was performed for electrophysiological studies (EPS), pacemaker indication, timing, and device interrogation for pacemaker dependency on follow up. A total of 24 patients received in-hospital PPM post-TAVR (14% of total cohort), and mean follow up was 22months. Indications for PPM included resting complete heart block (CHB; 15/24, 63%), left bundle branch block and abnormal electrophysiological study (EPS; 7/24, 29%), alternating bundle branch block (1/24, 4%) and tachy-brady syndrome (1/24, 4%). Pacemaker dependency (underlying ventricular asystole, complete heart block, or >50% pacing) occurred in 8/24 patients (33%) during follow-up, 7 of whom had resting CHB, and one with CHB invoked during EPS. Pacemaker dependency after TAVR is common among those that exhibited CHB, but not among those with a prolonged HV delay during EPS. Although preliminary, these observations are relevant to management of rhythm disturbances after TAVR, and may inform the practice of EPS-based PPM implantation. Copyright © 2017 Elsevier Inc. All rights reserved.

  2. A devices' game of thrones: cardiac resynchronization therapy vs. pacemaker.

    Science.gov (United States)

    Moura-Ferreira, Sara; Gonçalves, Helena; Oliveira, Marco; Primo, João; Fonseca, Paulo; Ribeiro, José; Santos, Elisabeth; Pelicano, Nuno; Martins, Dinis; Gama, Vasco

    2017-12-01

    Oversensing can interfere with biventricular pacing. Cardiac Resynchronization Therapy (CRT) output inhibition due to automatic brady mode change from a sensing to a pacing mode of a previously implanted pacemaker as it reached battery capacity depleted indicator has not been previously published in the medical literature. We report the first case of CRT output inhibition in a pacemaker dependent patient due to electrical stimuli from a previously right-sided implanted pacemaker, after unaware reversion of OVO mode (O = no chambers paced; V = ventricular sensing; O = no response to sensing) to backup VVI (V = ventricular pacing; V = ventricular sensing; I = inhibitory response to sensing) when it reached the elective replacement interval. This paper emphasizes the importance of knowing the distinct pacemaker brady mode behaviours after battery capacity depleted indicator has been reached, according to the pacemakers' manufacturer, including the possibility of automatic brady mode change from sensing to pacing mode. It also highlights the potential for severe bradycardia or asystole of this automatic brady mode change from a previously implanted pacemaker in pacemaker dependent patients submitted to CRT upgrade. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2017. For permissions, please email: journals.permissions@oup.com.

  3. Chromosomal abnormalities and copy number variations in fetal left-sided congenital heart defects.

    Science.gov (United States)

    Jansen, Fenna A R; Hoffer, Mariette J V; van Velzen, Christine L; Plati, Stephani Klingeman; Rijlaarsdam, Marry E B; Clur, Sally-Ann B; Blom, Nico A; Pajkrt, Eva; Bhola, Shama L; Knegt, Alida C; de Boer, Marion A; Haak, Monique C

    2016-02-01

    To demonstrate the spectrum of copy number variants (CNVs) in fetuses with isolated left-sided congenital heart defects (CHDs), and analyse genetic content. Between 2003 and 2012, 200 fetuses were identified with left-sided CHD. Exclusion criteria were chromosomal rearrangements, 22q11.2 microdeletion and/or extra-cardiac malformations (n = 64). We included cases with additional minor anomalies (n = 39), such as single umbilical artery. In 54 of 136 eligible cases, stored material was available for array analysis. CNVs were categorized as either (likely) benign, (likely) pathogenic or of unknown significance. In 18 of the 54 isolated left-sided CHDs we found 28 rare CNVs (prevalence 33%, average 1.6 CNV per person, size 10.6 kb-2.2 Mb). Our interpretation yielded clinically significant CNVs in two of 54 cases (4%) and variants of unknown significance in three other cases (6%). In left-sided CHDs that appear isolated, with normal chromosome analysis and 22q11.2 FISH analysis, array analysis detects clinically significant CNVs. When counselling parents of a fetus with a left-sided CHD it must be taken into consideration that aside from the cardiac characteristics, the presence of extra-cardiac malformations and chromosomal abnormalities influence the treatment plan and prognosis. © 2015 John Wiley & Sons, Ltd.

  4. A nationwide Danish cohort study challenging the categorisation into right-sided and left-sided colon cancer

    DEFF Research Database (Denmark)

    Jess, Per; Hansen, Iben Onsberg; Gamborg, Michael

    2013-01-01

    The categorisation of colon cancer (CC) into right-sided (RCC) and left-sided (LCC) disease may not capture more subtle variances in aetiology and prognosis. In a nationwide study, we investigated differences in clinical characteristics and survival of RCC versus LCC and of the complete range of CC...

  5. Minimally invasive radical pancreatectomy for left-sided pancreatic cancer: Current status and future perspectives

    Science.gov (United States)

    Kang, Chang Moo; Lee, Sung Hwan; Lee, Woo Jung

    2014-01-01

    Minimally invasive distal pancreatectomy with splenectomy has been regarded as a safe and effective treatment for benign and borderline malignant pancreatic lesions. However, its application for left-sided pancreatic cancer is still being debated. The clinical evidence for radical antegrade modular pancreatosplenectomy (RAMPS)-based minimally invasive approaches for left-sided pancreatic cancer was reviewed. Potential indications and surgical concepts for minimally invasive RAMPS were suggested. Despite the limited clinical evidence for minimally invasive distal pancreatectomy in left-sided pancreatic cancer, the currently available clinical evidence supports the use of laparoscopic distal pancreatectomy under oncologic principles in well-selected left sided pancreatic cancers. A pancreas-confined tumor with an intact fascia layer between the pancreas and left adrenal gland/kidney positioned more than 1 or 2 cm away from the celiac axis is thought to constitute a good condition for the use of margin-negative minimally invasive RAMPS. The use of minimally invasive (laparoscopic or robotic) anterior RAMPS is feasible and safe for margin-negative resection in well-selected left-sided pancreatic cancer. The oncologic feasibility of the procedure remains to be determined; however, the currently available interim results indicate that even oncologic outcomes will not be inferior to those of open radical distal pancreatosplenectomy. PMID:24605031

  6. Cardiac Pacemakers

    International Nuclear Information System (INIS)

    Fiandra, O.; Espasandin, W.; Fiandra, H.

    1984-01-01

    A complete survey of physiological biophysical,clinical and engineering aspects of cardiac facing,including the history and an assessment of possible future developments.Among the topics studied are: pacemakers, energy search, heart stimulating with pacemakers ,mathematical aspects of the electric cardio stimulation chronic, pacemaker implants,proceeding,treatment and control

  7. Difference between right and left side in total knee and unicondylar knee replacement: An interesting observation

    Directory of Open Access Journals (Sweden)

    Vasudevan Thirumal Selvan

    2005-02-01

    Full Text Available We report an observation made about the differences between right and left side in case of total knee and unicondylar knee replacement. It was found that unicondylar knee replacement was performed more commonly on the left side (66%, as compared to only 34% on right side, where as total knee replacement was more common on the right side (64% as compared to 36% on left side. The exact clinical utility of this difference is yet to be known.

  8. Surgical outcome of Fontan conversion and arrhythmia surgery: Need a pacemaker?

    Science.gov (United States)

    Terada, Takafumi; Sakurai, Hajime; Nonaka, Toshimichi; Sakurai, Takahisa; Sugiura, Junya; Taneichi, Tetsuyoshi; Ohtsuka, Ryohei

    2014-07-01

    Atrial tachyarrhythmias are frequent complications in the late period after the Fontan procedure, and important risk factors for a poor prognosis. The impact of Fontan conversion and arrhythmia surgery in failed Fontan patients has been described in many reports. We evaluated our experience with Fontan conversion procedures, concomitant arrhythmia surgery, and pacemaker implantation. We reviewed the hospital records of 25 consecutive patients who underwent a Fontan conversion procedure from January 2004 to March 2012. Twenty-four patients had arrhythmia surgery using cryoablation and radiofrequency ablation at the time of conversion. A bilateral atrial maze procedure was performed in 6 patients, right-side maze in 15, and isthmus block in 3. Three patients with a diagnosis of corrected transposition of the great arteries underwent simultaneous pacemaker implantation electively. There was no early death and one late death during a mean follow-up period of 21.2 months. Three tachyarrhythmia recurrences developed, and there were 4 occurrences of sinus bradycardia. Five of these patients required postoperative pacemaker implantation. The mid-term results of Fontan conversion and arrhythmia surgery in our institute were satisfactory. The occurrence of unexpected postoperative pacemaker requirement was high in the patients who underwent a right atrial or bilateral atrial maze procedure. Pacemaker or lead implantation is recommended for patients planned to undergo a right-side or full maze procedure. © The Author(s) 2013 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.

  9. Contribution of small conductance K+ channels to sinoatrial node pacemaker activity: insights from atrial-specific Na+ /Ca2+ exchange knockout mice.

    Science.gov (United States)

    Torrente, Angelo G; Zhang, Rui; Wang, Heidi; Zaini, Audrey; Kim, Brian; Yue, Xin; Philipson, Kenneth D; Goldhaber, Joshua I

    2017-06-15

    Repolarizing currents through K + channels are essential for proper sinoatrial node (SAN) pacemaking, but the influence of intracellular Ca 2+ on repolarization in the SAN is uncertain. We identified all three isoforms of Ca 2+ -activated small conductance K + (SK) channels in the murine SAN. SK channel blockade slows repolarization and subsequent depolarization of SAN cells. In the atrial-specific Na + /Ca 2+ exchanger (NCX) knockout mouse, cellular Ca 2+ accumulation during spontaneous SAN pacemaker activity produces intermittent hyperactivation of SK channels, leading to arrhythmic pauses alternating with bursts of pacing. These findings suggest that Ca 2+ -sensitive SK channels can translate changes in cellular Ca 2+ into a repolarizing current capable of modulating pacemaking. SK channels are a potential pharmacological target for modulating SAN rate or treating SAN dysfunction, particularly under conditions characterized by abnormal increases in diastolic Ca 2+ . Small conductance K + (SK) channels have been implicated as modulators of spontaneous depolarization and electrical conduction that may be involved in cardiac arrhythmia. However, neither their presence nor their contribution to sinoatrial node (SAN) pacemaker activity has been investigated. Using quantitative PCR (q-PCR), immunostaining and patch clamp recordings of membrane current and voltage, we identified all three SK isoforms (SK1, SK2 and SK3) in mouse SAN. Inhibition of SK channels with the specific blocker apamin prolonged action potentials (APs) in isolated SAN cells. Apamin also slowed diastolic depolarization and reduced pacemaker rate in isolated SAN cells and intact tissue. We investigated whether the Ca 2+ -sensitive nature of SK channels could explain arrhythmic SAN pacemaker activity in the atrial-specific Na + /Ca 2+ exchange (NCX) knockout (KO) mouse, a model of cellular Ca 2+ overload. SAN cells isolated from the NCX KO exhibited higher SK current than wildtype (WT) and apamin

  10. [Analysis of clinicopathologic and survival characteristics in patients with right-or left-sided colon cancer].

    Science.gov (United States)

    Hu, Junjie; Zhou, Zhixiang; Liang, Jianwei; Zhou, Haitao; Wang, Zheng; Zhang, Xingmao; Zeng, Weigen

    2015-07-28

    This study aimed to clarify the clinical and histological parameters, and survival difference between right- and left-sided colon cancer. We retrospectively analyzed the medical records (2006.1-2009.12) of 1 088 consecutive colon cancer patients who received surgery at our hospital. Right- and left-sided colon cancers were compared regarding the clinical and histological parameters. The survival analysis was performed by the Kaplan-Meier method, and the log-rank test was used to determine the statistical significance of differences. Right-sided colon cancer was associated with older age, a more advanced state, and poorly differentiated and undifferentiated adenocarcinoma (25.2% vs 13.2%), mucinous adenocarcinoma (33.5% vs 17.3%) and vascular invasion (9.9% vs 3.9%) were more commonly seen in right-sided colon cancer compared with right-sided colon cancer, and all these differences were statistically significant. Median overall survival was right, 67 months; and left, 68 months. The five-years overall survival of right- and left-sided colon cancer was I/II stage, 91.4% vs 88.6% (P = 0.819); III stage, 66.1% vs 75.4% (P = 0.010); and IV stage, 27.8% vs 38.5% (P = 0.020) respectively. Right- and left-sided colon cancers are significantly different regarding clinical and histological parameters. Right-sided colon cancers in stage III and IV have a worse prognosis.

  11. Persistent left superior vena cava: a case report and review of literature

    Directory of Open Access Journals (Sweden)

    Verma Gita

    2008-10-01

    Full Text Available Abstract Persistent left superior vena cava is rare but important congenital vascular anomaly. It results when the left superior cardinal vein caudal to the innominate vein fails to regress. It is most commonly observed in isolation but can be associated with other cardiovascular abnormalities including atrial septal defect, bicuspid aortic valve, coarctation of aorta, coronary sinus ostial atresia, and cor triatriatum. The presence of PLSVC can render access to the right side of heart challenging via the left subclavian approach, which is a common site of access utilized when placing pacemakers and Swan-Ganz catheters. Incidental notation of a dilated coronary sinus on echocardiography should raise the suspicion of PLSVC. The diagnosis should be confirmed by saline contrast echocardiography.

  12. Pacemaker and radiotherapy in breast cancer: is targeted intraoperative radiotherapy the answer in this setting?

    International Nuclear Information System (INIS)

    Keshtgar, Mohammed RS; Eaton, David J; Reynolds, Claire; Pigott, Katharine; Davidson, Tim; Gauter-Fleckenstein, Benjamin; Wenz, Frederik

    2012-01-01

    We present the case of an 83 year old woman with a cardiac pacemaker located close in distance to a subsequently diagnosed invasive ductal carcinoma of the left breast. Short range intraoperative radiotherapy was given following wide local excision and sentinel node biopsy. The challenges of using ionising radiation with pacemakers is also discussed

  13. Characterization of the Left-Sided Substrate in Arrhythmogenic Right Ventricular Cardiomyopathy.

    Science.gov (United States)

    Berte, Benjamin; Denis, Arnaud; Amraoui, Sana; Yamashita, Seigo; Komatsu, Yuki; Pillois, Xavier; Sacher, Frédéric; Mahida, Saagar; Wielandts, Jean-Yves; Sellal, Jean-Marc; Frontera, Antonio; Al Jefairi, Nora; Derval, Nicolas; Montaudon, Michel; Laurent, François; Hocini, Mélèze; Haïssaguerre, Michel; Jaïs, Pierre; Cochet, Hubert

    2015-12-01

    The correlates of left ventricular (LV) substrate in arrhythmogenic right ventricular (RV) cardiomyopathy are largely unknown. Thirty-two patients with arrhythmogenic RV cardiomyopathy (47±14 years; 6 women) were included. RV and LV dysplasia were defined from multidetector computed tomography and cardiac magnetic resonance imaging. Arrhythmias were characterized as right-sided or left-sided on 12-lead ECG recordings at baseline and during isoproterenol testing. In 14 patients, the imaging substrate was compared with voltage mapping and local abnormal ventricular activity. Imaging abnormalities were found in 32 (100%) and 21 (66%) patients on the RV and LV, respectively, intramyocardial fat on multidetector computed tomography being the most sensitive feature. LV involvement related to none of the Task Force criteria. Right-sided arrhythmias were more frequent than left-sided arrhythmias (P=0.003) although the latter were more frequent in case of LV involvement (P=0.02). The agreement between low voltage and fat on multidetector computed tomography was high on the RV when using either endocardial unipolar or epicardial bipolar data (κ=0.82 and κ=0.78, respectively) but lower on the LV (κ=0.54 for epicardial bipolar). LV local abnormal ventricular activity was found in all patients with LV involvement, and none of the others. The density of local abnormal ventricular activity within fat areas was similar between the RV and LV (P=0.57). LV substrate is frequent in arrhythmogenic RV cardiomyopathy, but poorly identified by current diagnostic strategies. Left-sided arrhythmias are more frequent in case of LV involvement. LV fat hosts the same density of local abnormal ventricular activity as RV fat, but is less efficiently detected by voltage mapping. These results support the need for alternative diagnostic strategies to identify LV dysplasia. © 2015 American Heart Association, Inc.

  14. Mathematical Models of Cardiac Pacemaking Function

    Science.gov (United States)

    Li, Pan; Lines, Glenn T.; Maleckar, Mary M.; Tveito, Aslak

    2013-10-01

    Over the past half century, there has been intense and fruitful interaction between experimental and computational investigations of cardiac function. This interaction has, for example, led to deep understanding of cardiac excitation-contraction coupling; how it works, as well as how it fails. However, many lines of inquiry remain unresolved, among them the initiation of each heartbeat. The sinoatrial node, a cluster of specialized pacemaking cells in the right atrium of the heart, spontaneously generates an electro-chemical wave that spreads through the atria and through the cardiac conduction system to the ventricles, initiating the contraction of cardiac muscle essential for pumping blood to the body. Despite the fundamental importance of this primary pacemaker, this process is still not fully understood, and ionic mechanisms underlying cardiac pacemaking function are currently under heated debate. Several mathematical models of sinoatrial node cell membrane electrophysiology have been constructed as based on different experimental data sets and hypotheses. As could be expected, these differing models offer diverse predictions about cardiac pacemaking activities. This paper aims to present the current state of debate over the origins of the pacemaking function of the sinoatrial node. Here, we will specifically review the state-of-the-art of cardiac pacemaker modeling, with a special emphasis on current discrepancies, limitations, and future challenges.

  15. Intraoperative colonic irrigation in the management of left sided ...

    African Journals Online (AJOL)

    Objectives: To evaluate the safety and benefits of antegrade intraoperative colonic irrigation (lavage) and primary anastomosis, after colonic resection, in the treatment of left sided large bowel emergencies. Design: A prospective descriptive study. Setting: Jos University Teaching Hospital, Jos, Nigeria. Participants: Thirty ...

  16. Monitoring the radiation dose to a multiprogrammable pacemaker during radical radiation therapy: A case report

    International Nuclear Information System (INIS)

    Muller-Runkel, R.; Orsolini, G.; Kalokhe, U.P.

    1990-01-01

    Multiprogrammable pacemakers, using complimentary metaloxide semiconductor (CMOS) circuitry, may fail during radiation therapy. We report about a patient who received 6,400 cGy for unresectable carcinoma of the left lung. In supine treatment position, arms raised above the head, the pacemaker was outside the treated area by a margin of at least 1 cm, shielded by cerrobend blocking mounted on a tray. From thermoluminescent dosimeter (TLD) measurements, we estimate that the pacemaker received 620 cGy in scatter doses. Its function was monitored before, during, and after completion of radiation therapy. The pacemaker was functioning normally until the patient's death 5 months after completion of treatment. The relevant electrocardiograms (ECGs) are presented

  17. Advanced Pacemaker

    Science.gov (United States)

    1990-01-01

    Synchrony, developed by St. Jude Medical's Cardiac Rhythm Management Division (formerly known as Pacesetter Systems, Inc.) is an advanced state-of-the-art implantable pacemaker that closely matches the natural rhythm of the heart. The companion element of the Synchrony Pacemaker System is the Programmer Analyzer APS-II which allows a doctor to reprogram and fine tune the pacemaker to each user's special requirements without surgery. The two-way communications capability that allows the physician to instruct and query the pacemaker is accomplished by bidirectional telemetry. APS-II features 28 pacing functions and thousands of programming combinations to accommodate diverse lifestyles. Microprocessor unit also records and stores pertinent patient data up to a year.

  18. A missing link between RON expression and oncological outcomes in resected left-sided pancreatic cancer.

    Science.gov (United States)

    Han, Dai Hoon; Kang, Chang Moo; Lee, Sung Whan; Hwang, Ho Kyoung; Lee, Woo Jung

    2017-10-01

    Alteration and activation of recepteur d'origine nantais (RON) expression is known to be associated with cancer progression and decreased survival in various types of human cancer, including pancreatic cancer. Therefore, in the present study, RON expression levels were determined in resected left-sided pancreatic cancer to evaluate the potential oncological role of RON in the clinical setting of distal pancreatic cancer. From January 2005 to December 2011, a total of 57 patients underwent radical distal pancreatectomy for left-sided pancreatic cancer. Ductal adenocarcinoma was confirmed in all patients. Among these patients, 17 patients who received preoperative neoadjuvant treatment and 7 patients without available paraffin-embedded tissue blocks were excluded from the present study. RON expression in a the pancreatic cancer cell lines ASPC-1, BxPC-3, MiaPaCa-3 and Panc-1, as well as in resected left-sided pancreatic cancer specimens was determined by Western blot analysis. RON and vascular endothelial growth factor (VEGF) overexpression in resected left-sided pancreatic cancer was also evaluated by immunohistochemistry using pre-diluted anti-RON and anti-VEGF antibodies. An association was identified between the oncological outcome and RON overexpression. Increased levels of RON expression were observed in two pancreatic cancer cell lines, AsPC-1 and BxPC-3. RON overexpression was detected in specimens from 15/33 patients (45.5%) using immunohistochemistry. No significant association was identified between RON overexpression and VEGF overexpression (25.5 vs. 87.9%; P=0.667). No significant differences in disease-free survival or disease-specific survival associated with RON overexpression were identified. Although the results of previous studies have suggested that RON is a potential target for the treatment of pancreatic cancer, in the present study no association between RON overexpression and any adverse oncological effect was identified.

  19. Emergency surgery for obstructing colonic cancer: a comparison between right-sided and left-sided lesions.

    Science.gov (United States)

    Faucheron, J-L; Paquette, B; Trilling, B; Heyd, B; Koch, S; Mantion, G

    2018-02-01

    Few studies compare management and outcomes of obstructive colonic cancer, depending on the tumor site. We aim to evaluate the differences in patient characteristics, tumor characteristics, and outcomes of emergency surgery for obstructive right-sided versus left-sided colonic cancers. Between 2000 and 2009, 71 consecutive patients had an emergency colectomy following strict and clear definition of obstruction in a single institution. We retrospectively analyzed pre, per, and postoperative data that were prospectively collected. There were 31 and 40 patients in the right and left group, respectively. Patients aged over 80 were more frequent in the right group (p = 0.03). At operation, ileocecal valve was less often competent in the right group (p = 0.03). The one-stage strategy was more frequent in the right group (p = 0.008). Patients in the right group had a higher rate of nodes invasion (p = 0.04). One- and two-year mortality rate in the right group had a tendency to be higher. Patients presenting with a right obstructive colonic cancer are older, have a more advanced locoregional disease, and are more often treated in a one-stage strategy than patients with a left obstructive tumor.

  20. Zygotic LvBMP5-8 is required for skeletal patterning and for left-right but not dorsal-ventral specification in the sea urchin embryo.

    Science.gov (United States)

    Piacentino, Michael L; Chung, Oliver; Ramachandran, Janani; Zuch, Daniel T; Yu, Jia; Conaway, Evan A; Reyna, Arlene E; Bradham, Cynthia A

    2016-04-01

    Skeletal patterning in the sea urchin embryo requires coordinated signaling between the pattern-dictating ectoderm and the skeletogenic primary mesenchyme cells (PMCs); recent studies have begun to uncover the molecular basis for this process. Using an unbiased RNA-Seq-based screen, we have previously identified the TGF-ß superfamily ligand, LvBMP5-8, as a skeletal patterning gene in Lytechinus variegatus embryos. This result is surprising, since both BMP5-8 and BMP2/4 ligands have been implicated in sea urchin dorsal-ventral (DV) and left-right (LR) axis specification. Here, we demonstrate that zygotic LvBMP5-8 is required for normal skeletal patterning on the left side, as well as for normal PMC positioning during gastrulation. Zygotic LvBMP5-8 is required for expression of the left-side marker soxE, suggesting that LvBMP5-8 is required for left-side specification. Interestingly, we also find that LvBMP5-8 knockdown suppresses serotonergic neurogenesis on the left side. While LvBMP5-8 overexpression is sufficient to dorsalize embryos, we find that zygotic LvBMP5-8 is not required for normal DV specification or development. In addition, ectopic LvBMP5-8 does not dorsalize LvBMP2/4 morphant embryos, indicating that, in the absence of BMP2/4, BMP5-8 is insufficient to specify dorsal. Taken together, our data demonstrate that zygotic LvBMP5-8 signaling is essential for left-side specification, and for normal left-side skeletal and neural patterning, but not for DV specification. Thus, while both BMP2/4 and BMP5-8 regulate LR axis specification, BMP2/4 but not zygotic BMP5-8 regulates DV axis specification in sea urchin embryos. Copyright © 2016 Elsevier Inc. All rights reserved.

  1. Mathematical Models of Cardiac Pacemaking Function

    Directory of Open Access Journals (Sweden)

    Pan eLi

    2013-10-01

    Full Text Available Over the past half century, there has been intense and fruitful interaction between experimental and computational investigations of cardiac function. This interaction has, for example, led to deep understanding of cardiac excitation-contraction coupling; how it works, as well as how it fails. However, many lines of inquiry remain unresolved, among them the initiation of each heartbeat. The sinoatrial node, a cluster of specialized pacemaking cells in the right atrium of the heart, spontaneously generates an electro-chemical wave that spreads through the atria and through the cardiac conduction system to the ventricles, initiating the contraction of cardiac muscle essential for pumping blood to the body. Despite the fundamental importance of this primary pacemaker, this process is still not fully understood, and ionic mechanisms underlying cardiac pacemaking function are currently under heated debate. Several mathematical models of sinoatrial node cell membrane electrophysiology have been constructed as based on different experimental data sets and hypotheses. As could be expected, these differing models offer diverse predictions about cardiac pacemaking activities. This paper aims to present the current state of debate over the origins of the pacemaking function of the sinoatrial node. Here, we will specifically review the state-of-the-art of cardiac pacemaker modeling, with a special emphasis on current discrepancies, limitations, and future challenges.

  2. Ductal carcinoma of the breast in the pacemaker generator's pocket.

    Science.gov (United States)

    Zonca, P; Herokova, J; Cambal, M; Jacobi, C A

    2009-01-01

    Authors present a case of a 78-year-old female patient with invasive ductal adenocarcinoma in the pacemaker, s pocket. A decubitus-like tumor had developed in this place, and has been missinterpretated as a benign lesion for 5 months. Diagnosis was done with a time delay. An excisional biopsy revealed annvasive ductal adenocarcinoma. The first step was the implantation of a new pacemaker generator performed on the opposite side. The second step was a modified radical mastectomy, according to Madden, and the removal of the originally implanted pacemaker generator. Radiotherapy and hormonal adjuvant therapy were applied after surgery. The patient was followed-up at an out-patient clinic, and died 25 months after diagnosis because of generalization of the disease (Fig. 2, Ref. 35). Full Text (Free, PDF) www.bmj.sk.

  3. Right Site, Wrong Route - Cannulating the Left Internal Jugular Vein.

    Science.gov (United States)

    Paik, Peter; Arukala, Sanjay K; Sule, Anupam A

    2018-01-09

    Central venous catheters are placed in approximately five million patients annually in the US. The preferred site of insertion is one with fewer risks and easier access. Although the right internal jugular vein is preferred, on occasion, the left internal jugular may have to be accessed. A patient was admitted for septic shock, cerebrovascular accident, and non-ST-segment elevation myocardial infarction. A central venous line was needed for antibiotic and vasopressor administration. Due to trauma from a fall to the right side and previously failed catheterization attempts at the left subclavian and femoral veins, the left internal jugular vein was accessed. On chest radiography for confirmation, the left internal jugular central venous catheter was seen projecting down the left paraspinal region. It did not take the expected course across the midline toward the right and into the superior vena cava (SVC). A review of a computed tomography (CT) scan of the chest with contrast done on a prior admission revealed a duplicated SVC on the left side that had not been reported in the original CT scan interpretation. A left-sided SVC is present in approximately 0.3% to 0.5% of the population, with 90% of these draining into the coronary sinus. During placements of central venous lines and pacemakers, irritation of the coronary sinus may result in hypotension, arrhythmia, myocardial ischemia, or cardiac arrest. A widened mediastinum can be an indication of a duplicated SVC. When attempting a left internal jugular vein central venous catheter placement, it is important to be aware of venous anomalies in order to prevent complications.

  4. RIGHT-SIDED AORTIC ARCH WITH ABERRANT LEFT SUBCLAVIAN ARTERY AND DUPLICATION OF SUPERIOR VENA CAVA

    Directory of Open Access Journals (Sweden)

    Parikhita Hazarika

    2017-08-01

    Full Text Available Right-sided aortic arch is a rare anatomical variant present in about 0.1% of the adult population.1,2 Half of the cases are associated with an aberrant left subclavian artery (0.05%-0.1%. Right-sided aortic arch with aberrant left subclavian artery is less common than left-sided aortic arch with aberrant right subclavian artery (0.5-2.0%.3,4 A rightsided aortic arch is an anatomic variant resulting from persistence of the right fourth aortic arch and involution of the left. It can be associated with an aberrant left subclavian artery arises from Kommerell’s diverticulum. It is usually asymptomatic and diagnosed incidentally during adult age. A 40-year-old male presented with cough and a hump in the back. The patient was evaluated for scoliosis and plain CT thorax was done.

  5. Right-sided aortic arch with anomalous origin of the left subclavian artery: Case report

    Directory of Open Access Journals (Sweden)

    Vučurević Goran

    2011-01-01

    Full Text Available Introduction. A right-sided aortic arch is a rare congenital defect of the aorta with incidence of 0.05% to 0.1% reported in published series. Usually it is associated with congenital heart anomalies and esophageal and tracheal compression symptoms. We present a case of a right-sided aortic arch of anomalous left subclavian artery origin, accidentally revealed during multislice CT (MSCT supraaortic branches angiography. Case Outline. A 53-year-old female patient was examined at the Outpatients’ Unit of the Vascular Surgery University Clinic for vertigo, occasional dizziness and difficulty with swallowing. Physical examination revealed a murmur of the left supraclavicular space, with 15 mmHg lower rate of arterial tension on the left arm. Ultrasound of carotid arteries revealed 60% stenosis of the left subclavian artery and bilateral internal carotid artery elongation. MSCT angiography revealed a right-sided aortic arch with aberrant separation of the left subclavian artery that was narrowed 50%, while internal carotid arteries were marginally elongated. There was no need for surgical treatment or percutaneous interventions, so that conservative treatment was indicated. Conclusion. A right-sided aortic arch is a very rare anomaly of the location and branching of the aorta. Multislice CT angiography is of great importance in the diagnostics of this rare disease.

  6. Detection of Left-Sided and Right-Sided Hearing Loss via Fractional Fourier Transform

    Directory of Open Access Journals (Sweden)

    Shuihua Wang

    2016-05-01

    Full Text Available In order to detect hearing loss more efficiently and accurately, this study proposed a new method based on fractional Fourier transform (FRFT. Three-dimensional volumetric magnetic resonance images were obtained from 15 patients with left-sided hearing loss (LHL, 20 healthy controls (HC, and 14 patients with right-sided hearing loss (RHL. Twenty-five FRFT spectrums were reduced by principal component analysis with thresholds of 90%, 95%, and 98%, respectively. The classifier is the single-hidden-layer feed-forward neural network (SFN trained by the Levenberg–Marquardt algorithm. The results showed that the accuracies of all three classes are higher than 95%. In all, our method is promising and may raise interest from other researchers.

  7. Novel approach to epicardial pacemaker implantation in patients with limited venous access.

    Science.gov (United States)

    Costa, Roberto; Scanavacca, Mauricio; da Silva, Kátia Regina; Martinelli Filho, Martino; Carrillo, Roger

    2013-11-01

    Limited venous access in certain patients increases the procedural risk and complexity of conventional transvenous pacemaker implantation. The purpose of this study was to determine a minimally invasive epicardial approach using pericardial reflections for dual-chamber pacemaker implantation in patients with limited venous access. Between June 2006 and November 2011, 15 patients underwent epicardial pacemaker implantation. Procedures were performed through a minimally invasive subxiphoid approach and pericardial window with subsequent fluoroscopy-assisted lead placement. Mean patient age was 46.4 ± 15.3 years (9 male [(60.0%], 6 female [40.0%]). The new surgical approach was used in patients determined to have limited venous access due to multiple abandoned leads in 5 (33.3%), venous occlusion in 3 (20.0%), intravascular retention of lead fragments from prior extraction in 3 (20.0%), tricuspid valve vegetation currently under treatment in 2 (13.3%), and unrepaired intracardiac defects in 2 (13.3%). All procedures were successful with no perioperative complications or early deaths. Mean operating time for isolated pacemaker implantation was 231.7 ± 33.5 minutes. Lead placement on the superior aspect of right atrium, through the transverse sinus, was possible in 12 patients. In the remaining 3 patients, the atrial lead was implanted on the left atrium through the oblique sinus, the postcaval recess, or the left pulmonary vein recess. None of the patients displayed pacing or sensing dysfunction, and all parameters remained stable throughout the follow-up period of 36.8 ± 25.1 months. Epicardial pacemaker implantation through pericardial reflections is an effective alternative therapy for those patients requiring physiologic pacing in whom venous access is limited. © 2013 Heart Rhythm Society. All rights reserved.

  8. Modern Perspectives on Numerical Modeling of Cardiac Pacemaker Cell

    Science.gov (United States)

    Maltsev, Victor A.; Yaniv, Yael; Maltsev, Anna V.; Stern, Michael D.; Lakatta, Edward G.

    2015-01-01

    Cardiac pacemaking is a complex phenomenon that is still not completely understood. Together with experimental studies, numerical modeling has been traditionally used to acquire mechanistic insights in this research area. This review summarizes the present state of numerical modeling of the cardiac pacemaker, including approaches to resolve present paradoxes and controversies. Specifically we discuss the requirement for realistic modeling to consider symmetrical importance of both intracellular and cell membrane processes (within a recent “coupled-clock” theory). Promising future developments of the complex pacemaker system models include the introduction of local calcium control, mitochondria function, and biochemical regulation of protein phosphorylation and cAMP production. Modern numerical and theoretical methods such as multi-parameter sensitivity analyses within extended populations of models and bifurcation analyses are also important for the definition of the most realistic parameters that describe a robust, yet simultaneously flexible operation of the coupled-clock pacemaker cell system. The systems approach to exploring cardiac pacemaker function will guide development of new therapies, such as biological pacemakers for treating insufficient cardiac pacemaker function that becomes especially prevalent with advancing age. PMID:24748434

  9. Pacemaker (image)

    Science.gov (United States)

    A pacemaker is a small, battery-operated electronic device which is inserted under the skin to help the heart beat regularly and at an appropriate rate. The pacemaker has leads that travel through a large vein ...

  10. Echocardiography-guided Radiofrequency Catheter Ablation of Atrioventricular Node and VVI Pacemaker Implantation

    Directory of Open Access Journals (Sweden)

    T Guo

    2014-05-01

    Full Text Available Objective: This study is to evaluate the feasibility and safety of intracardiac radiofrequency catheter ablation (RFCA of the atrioventricular node (AVN and pacemaker implantation using transthoracic echocardiography. Methods: Eleven patients – six males and five females (mean age 66 years – with persistent or permanent atrial fibrillation/atrial flutter received RFCA of AVN and VVI pacemaker implantation (paces and senses the ventricle and is inhibited if it senses ventricular activity. Under transthoracic echocardiography, the electrode catheters were positioned intracardiac, and target ablation was performed, with the permanent pacemaking catheter in the left subclavian vein and the ablation catheter in the right femoral vein. The multi-view imaging and dynamic observation applied during the stable AV dissociation were successful. Results: Atrioventricular node ablation and permanent pacemaker implantation in 11 patients were completed successfully without X-ray exposure. The operation success rate was 100%. All patients recovered well within the follow-up period. Conclusions: Radiofrequency catheter ablation of AVN and VVI pacemaker implantation under transthoracic echocardiography guidance is a safe, easy and feasible approach. This procedure could be an important supplemental measure to catheter ablation of arrhythmia under routine X-ray fluoroscopy.

  11. Factors affecting colonoscope insertion time in patients with or without a colostomy after left-sided colorectal resection.

    Science.gov (United States)

    Jang, Hui Won; Kim, Yoon Nam; Nam, Chung Mo; Lee, Hyun Jung; Park, Soo Jung; Hong, Sung Pil; Kim, Tae Il; Kim, Won Ho; Cheon, Jae Hee

    2012-12-01

    We examined whether the insertion time for colonoscopies performed after left-sided resection was different in patients with a colostomy from that in patients without a colostomy and identified factors that could impact colonoscopy performance. We included consecutive patients who underwent colonoscopy between July 2005 and March 2011 after left-sided colorectal resection for colorectal cancer. We classified surgical methods according to the presence or absence of a colostomy and evaluated colonoscope insertion time retrospectively. Furthermore, we analyzed factors that might affect insertion time. A total of 1,041 patients underwent colonoscopy after left-sided colorectal resection during the study period. The colonoscopy completion rate was 98.6 %, and the mean insertion time was 6.1 ± 4.6 min (median 4.7 min, range 0.3-35.8 min). A shorter resection length of colon, the presence of a colostomy, and a lower endoscopist case volume were found to be independent factors associated with prolonged insertion time in patients with left-sided colorectal resection. Among experienced colonoscopists, no colonoscopy-associated or clinical factors were found to affect insertion time. However, a shorter resection length of colon, the presence of a colostomy, and poor bowel preparation were associated with prolonged insertion time among inexperienced endoscopists. We identified three factors that affect colonoscope insertion time after left-sided colorectal resection, including the presence of a colostomy. Inexperienced endoscopists were much more affected by the presence of a colostomy after left-sided colorectal resection. These findings have implications for the practice and teaching of colonoscopy after left-sided colorectal resection.

  12. US-guided percutaneous transhepatic biliary drainage: comparative study of right-sided and left-sided approach

    International Nuclear Information System (INIS)

    Kim, Young Hwan; Cha, Soon Joo

    2002-01-01

    To compare the feasibility and safety of US-guided right and left percutaneous transhepatic biliary drainage (PTBD). Between March 1998 and May 1999, 32 patients underwent 36 US-guided right or left PTBD in referred order, alternatively. The causes of biliary obstruction were bile duct stone (n=2), bile duct carcinoma (n=10), carcinoma of the pancreas (n=9), GB carcinoma (n=7), metastasis to the porta hepatis (n=3), and carcinoma of the ampulla of vater (n=1). Technical success, procedure time, fluoroscopic time, and complications were evaluated. PTBD was successful in 94% of both right and left approach. The average procedure time was 9.7 ±3.8 min. in the right approach and 9.6 ±3.1 min. in the left approach, respectively (p=0.794). The average fluoroscopic time were 3.9±2.4 min. in the right approach and 3.8±2.2 min. in the left approach (p=0.892). A major complication, bile peritonitis, occurred in one of 16 patient with right-sided approach. Minor complications occurred in six right (2 hemobilia, 3 tube malfunction, 1 cholangitis) and three left (1 hemobilia, 1 fever, 1cholangitis) PTBD. There were no significant difference in the complication rates between right and left PTBD (p=0.729). There were no significant differences in feasibility and safety in US-guided right and left PTBD

  13. US-guided percutaneous transhepatic biliary drainage: comparative study of right-sided and left-sided approach

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Young Hwan; Cha, Soon Joo [College of Medicine, Inje Univ., Kimhae (Korea, Republic of)

    2002-02-01

    To compare the feasibility and safety of US-guided right and left percutaneous transhepatic biliary drainage (PTBD). Between March 1998 and May 1999, 32 patients underwent 36 US-guided right or left PTBD in referred order, alternatively. The causes of biliary obstruction were bile duct stone (n=2), bile duct carcinoma (n=10), carcinoma of the pancreas (n=9), GB carcinoma (n=7), metastasis to the porta hepatis (n=3), and carcinoma of the ampulla of vater (n=1). Technical success, procedure time, fluoroscopic time, and complications were evaluated. PTBD was successful in 94% of both right and left approach. The average procedure time was 9.7 {+-}3.8 min. in the right approach and 9.6 {+-}3.1 min. in the left approach, respectively (p=0.794). The average fluoroscopic time were 3.9{+-}2.4 min. in the right approach and 3.8{+-}2.2 min. in the left approach (p=0.892). A major complication, bile peritonitis, occurred in one of 16 patient with right-sided approach. Minor complications occurred in six right (2 hemobilia, 3 tube malfunction, 1 cholangitis) and three left (1 hemobilia, 1 fever, 1cholangitis) PTBD. There were no significant difference in the complication rates between right and left PTBD (p=0.729). There were no significant differences in feasibility and safety in US-guided right and left PTBD.

  14. Assessment of Intrafraction Breathing Motion on Left Anterior Descending Artery Dose During Left-Sided Breast Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    El-Sherif, Omar, E-mail: Omar.ElSherif@lhsc.on.ca [Department of Medical Biophysics, University of Western Ontario, London, Ontario (Canada); Department of Physics, London Regional Cancer Program, London, Ontario (Canada); Yu, Edward [Department of Radiation Oncology, London Regional Cancer Program, London, Ontario (Canada); Xhaferllari, Ilma [Department of Medical Biophysics, University of Western Ontario, London, Ontario (Canada); Department of Physics, London Regional Cancer Program, London, Ontario (Canada); Gaede, Stewart [Department of Medical Biophysics, University of Western Ontario, London, Ontario (Canada); Department of Physics, London Regional Cancer Program, London, Ontario (Canada); Department of Radiation Oncology, London Regional Cancer Program, London, Ontario (Canada)

    2016-07-01

    Purpose: To use 4-dimensional computed tomography (4D-CT) imaging to predict the level of uncertainty in cardiac dose estimates of the left anterior descending artery that arises due to breathing motion during radiation therapy for left-sided breast cancer. Methods and Materials: The fast helical CT (FH-CT) and 4D-CT of 30 left-sided breast cancer patients were retrospectively analyzed. Treatment plans were created on the FH-CT. The original treatment plan was then superimposed onto all 10 phases of the 4D-CT to quantify the dosimetric impact of respiratory motion through 4D dose accumulation (4D-dose). Dose-volume histograms for the heart, left ventricle (LV), and left anterior descending (LAD) artery obtained from the FH-CT were compared with those obtained from the 4D-dose. Results: The 95% confidence interval of 4D-dose and FH-CT differences in mean dose estimates for the heart, LV, and LAD were ±0.5 Gy, ±1.0 Gy, and ±8.7 Gy, respectively. Conclusion: Fast helical CT is a good approximation for doses to the heart and LV; however, dose estimates for the LAD are susceptible to uncertainties that arise due to intrafraction breathing motion that cannot be ascertained without the additional information obtained from 4D-CT and dose accumulation. For future clinical studies, we suggest the use of 4D-CT–derived dose-volume histograms for estimating the dose to the LAD.

  15. No inherent left and right side in human 'mental number line': evidence from right brain damage.

    Science.gov (United States)

    Aiello, Marilena; Jacquin-Courtois, Sophie; Merola, Sheila; Ottaviani, Teresa; Tomaiuolo, Francesco; Bueti, Domenica; Rossetti, Yves; Doricchi, Fabrizio

    2012-08-01

    Spatial reasoning has a relevant role in mathematics and helps daily computational activities. It is widely assumed that in cultures with left-to-right reading, numbers are organized along the mental equivalent of a ruler, the mental number line, with small magnitudes located to the left of larger ones. Patients with right brain damage can disregard smaller numbers while mentally setting the midpoint of number intervals. This has been interpreted as a sign of spatial neglect for numbers on the left side of the mental number line and taken as a strong argument for the intrinsic left-to-right organization of the mental number line. Here, we put forward the understanding of this cognitive disability by discovering that patients with right brain damage disregard smaller numbers both when these are mapped on the left side of the mental number line and on the right side of an imagined clock face. This shows that the right hemisphere supports the representation of small numerical magnitudes independently from their mapping on the left or the right side of a spatial-mental layout. In addition, the study of the anatomical correlates through voxel-based lesion-symptom mapping and the mapping of lesion peaks on the diffusion tensor imaging-based reconstruction of white matter pathways showed that the rightward bias in the imagined clock-face was correlated with lesions of high-level middle temporal visual areas that code stimuli in object-centred spatial coordinates, i.e. stimuli that, like a clock face, have an inherent left and right side. In contrast, bias towards higher numbers on the mental number line was linked to white matter damage in the frontal component of the parietal-frontal number network. These anatomical findings show that the human brain does not represent the mental number line as an object with an inherent left and right side. We conclude that the bias towards higher numbers in the mental bisection of number intervals does not depend on left side spatial

  16. Complicated left-sided native valve endocarditis in adults: risk classification for mortality.

    Science.gov (United States)

    Hasbun, Rodrigo; Vikram, Holenarasipur R; Barakat, Lydia A; Buenconsejo, Joan; Quagliarello, Vincent J

    2003-04-16

    Complicated left-sided native valve endocarditis causes significant morbidity and mortality in adults. Lack of valid data regarding estimation of prognosis makes management of this condition difficult. To derive and externally validate a prognostic classification system for adults with complicated left-sided native valve endocarditis. Retrospective observational cohort study conducted from January 1990 to January 2000 at 7 Connecticut hospitals among 513 patients older than 16 years who experienced complicated left-sided native valve endocarditis and who were divided into derivation (n = 259) and validation (n = 254) cohorts. All-cause mortality at 6 months after baseline. In the derivation and validation cohorts, the 6-month mortality rates were 25% and 26%, respectively. Five baseline features were independently associated with 6-month mortality (comorbidity [P =.03], abnormal mental status [P =.02], moderate to severe congestive heart failure [P =.01], bacterial etiology other than viridans streptococci [Pclassification system. In the derivation cohort, patients were classified into 4 groups with increasing risk for 6-month mortality: 5%, 15%, 31%, and 59% (Pendocarditis can be accurately risk stratified using baseline features into 4 groups of prognostic severity. This prognostic classification system might be useful for facilitating management decisions.

  17. Left-sided and duplicate inferior vena cava: a case series and review.

    Science.gov (United States)

    Ang, Wee Choen; Doyle, Terry; Stringer, Mark D

    2013-11-01

    Left-sided and duplicate inferior vena cava (IVC) are two major anatomical variants within the spectrum of IVC malformations, both of which are developmental abnormalities of the supracardinal veins. Four clinical cases are described to highlight the computed tomographic appearances of these vascular malformations and provide novel data on venous dimensions. A systematic review of the recent literature (2000-2011) was conducted focusing on the anatomy, demographics, and associated pathology (congenital and acquired) of isolated left-sided and duplicate IVC. A total of 73 relevant articles were retrieved, consisting of case reports and small case series. The prevalence of left-sided IVC is about 0.1-0.4% and that for duplicate IVC about 0.3-0.4%; both anomalies show a slight male preponderance. In each condition, there are documented variations in the course and tributaries of the IVC. The clinical importance of these anomalies lies in three principal areas: the potential for misdiagnosis on imaging; technical difficulties during retroperitoneal surgery (particularly abdominal aortic aneurysm repair and live donor nephrectomy); and their significance in relation to the etiology and management of venous thromboembolism. Copyright © 2012 Wiley Periodicals, Inc.

  18. Influencing programmable pacemakers by radiation therapy

    International Nuclear Information System (INIS)

    Wilm, M.; Kronholz, H.L.; Schuetz, J.; Koch, T.

    1994-01-01

    More than 300,000 pacemakers are implanted worldwide. During radiation therapy a damage of the pacemaker elektronic is possible. Twenty pacemakers have been irradiated with photons or electrons experimentally in three different situations: a) pacemaker and pacemaker electrode outside of the irradiation field; b) pacemaker outside, pacemaker electrode inside the irradiation field; c) all things inside the irradiation field. The voltage in the pacemaker electrode produced by the electric field of the accelerator did not exceed 0.8 mV if the electrode was outside the irradiation field. Induced voltage was up to 1.2 mV during irradiation with electrons (18 MeV) and the electrode being inside the treatment field with more than two thirds of its length. After delivering of not more than 10 Gy (photons) to the pacemaker, a decreasing amplitude of the pacemaker pulse occurred. The pulse frequency did not show any deviation. This seems to signal a severe early irreversible damage of the pacemaker that may cause sudden breakdown days or weeks after radiation. Two pacemakers showed a complete breakdown after irradiation with not more than 10 Gy. The others had a complete breakdown beyond doses of 50 Gy. (orig./MG) [de

  19. Insight into the da Vinci® Xi - technical notes for single-docking left-sided colorectal procedures.

    Science.gov (United States)

    Ngu, James Chi-Yong; Sim, Sarah; Yusof, Sulaiman; Ng, Chee-Yung; Wong, Andrew Siang-Yih

    2017-12-01

    The adoption of robot-assisted laparoscopic colorectal surgery has been hampered by issues with docking, operative duration, technical difficulties in multi-quadrant access, and cost. The da Vinci® Xi has been designed to overcome some of these limitations. We describe our experience with the system and offer technical insights to its application in left-sided colorectal procedures. Our initial series of left-sided robotic colorectal procedures was evaluated. Patient demographics and operative outcomes were recorded prospectively using a predefined database. Between March 2015 and April 2016, 54 cases of robot-assisted laparoscopic left-sided colorectal procedures were successfully completed with no cases of conversion. The majority were low anterior resections for colorectal malignancies. Using the da Vinci® Xi Surgical System, multi-quadrant surgery involving dissection from the splenic flexure to the pelvis was possible without redocking. The da Vinci® Xi simplifies the docking procedure and makes single-docking feasible for multi-quadrant left-sided colorectal procedures. Copyright © 2016 John Wiley & Sons, Ltd.

  20. Silent Atrial Fibrillation in Elderly Pacemaker Users: A Randomized Trial Using Home Monitoring.

    Science.gov (United States)

    Lima, Ceb; Martinelli, M; Peixoto, G L; Siqueira, S F; Wajngarten, Maurício; Silva, Rodrigo Tavares; Costa, Roberto; Filho, Roberto; Ramires, José Antônio Franchini

    2016-05-01

    Pacemaker with remote monitoring (PRM) may be useful for silent atrial fibrillation (AF) detection. The aims of this study were to evaluate the incidence of silent AF, the role of PRM, and to determine predictors of silent AF occurrence. Three hundred elderly patients with permanent pacemaker (PPM) were randomly assigned to the remote group (RG) or control group (CG). All patients received PPM with remote monitoring capabilities. Primary end point was AF occurrence rate and the secondary end points were time to AF detection and number of days with AF. During the average follow-up of 15.7±7.7 months, AF episodes were detected in 21.6% (RG = 24% vs CG = 19.3%, P = 0.36]. There was no difference in the time to detect the first AF episode. However, the median time to detect AF recurrence in the RG was lower than that in the CG (54 days vs 100 days, P = 0.004). The average number of days with AF was 16.0 and 51.2 in the RG and CG, respectively (P = 0.028). Predictors of silent AF were left atrial diameter (odds ratio [OR] 1.2; 95% CI = 1.1-1.3; P pacemaker; left atrial diameter and diastolic dysfunction were predictors of its occurrence. AF monitoring by means of pacemaker is a valuable tool for silent AF detection and continuous remote monitoring allows early AF recurrence detection and reduces the number of days with AF. © 2015 Wiley Periodicals, Inc.

  1. Infection of the left sided gallbladder simulating liver abscess and gastric antral tumor

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Jong Woo; Ahn, Jae Sung [St Mary' s Hospital Catholic Medical College, Seoul (Korea, Republic of)

    1972-12-15

    The left sided gallbladder, being a rare occurrence, may create diagnostic problem especially when it becomes infected. Recently, we have experienced 2 cases of infection of left sided gallbladder at the Department of Radiology, St Mary's Hospital, Catholic Medical College. Case I was a 32-year-old housewife with dyspepsia, epigastric pain and recurring bouts of high fever of 2 months duration. Under the provisional diagnosis of liver abscess liver scan was performed which revealed well defined cold area in the left lobe supporting the provisional diagnosis. Tender cystic mass was aspirated and pus was drained. The injected air at the occasion of aspiration outlined well defined cystic structure. Because aspiration failed to control the symptoms, laparotomy was performed. Surgeon found a left sided gallbladder which was empyematous. Case II was a 38 year old business man admitted to the hospital because of tender ill defined mass in mid epigastrium. The patient was febrile (39.4 .deg. C) and anemic. An upper GI series disclosed an irregular filling defect involving the antrum of the stomach. The mass was primarily extrinsic but was closely fixed to the gastric wall. Surgical intervention revealed chronic cholecystitis and pericholecystitis with thick adhesion to the distal stomach.

  2. Pacemaker Dependency after Cardiac Surgery: A Systematic Review of Current Evidence.

    Science.gov (United States)

    Steyers, Curtis M; Khera, Rohan; Bhave, Prashant

    2015-01-01

    Severe postoperative conduction disturbances requiring permanent pacemaker implantation frequently occur following cardiac surgery. Little is known about the long-term pacing requirements and risk factors for pacemaker dependency in this population. We performed a systematic review of the literature addressing rates and predictors of pacemaker dependency in patients requiring permanent pacemaker implantation after cardiac surgery. Using a comprehensive search of the Medline, Web of Science and EMBASE databases, studies were selected for review based on predetermined inclusion and exclusion criteria. A total of 8 studies addressing the endpoint of pacemaker-dependency were identified, while 3 studies were found that addressed the recovery of atrioventricular (AV) conduction endpoint. There were 10 unique studies with a total of 780 patients. Mean follow-up ranged from 6-72 months. Pacemaker dependency rates ranged from 32%-91% and recovery of AV conduction ranged from 16%-42%. There was significant heterogeneity with respect to the definition of pacemaker dependency. Several patient and procedure-specific variables were found to be independently associated with pacemaker dependency, but these were not consistent between studies. Pacemaker dependency following cardiac surgery occurs with variable frequency. While individual studies have identified various perioperative risk factors for pacemaker dependency and non-resolution of AV conduction disease, results have been inconsistent. Well-conducted studies using a uniform definition of pacemaker dependency might identify patients who will benefit most from early permanent pacemaker implantation after cardiac surgery.

  3. Case Report of Radiotherapy to a Breast Cancer Patient with a Pacemaker

    Energy Technology Data Exchange (ETDEWEB)

    Chae, Seung Hoon; Park, Jang Pil; Lee, Yang Hoon; Yoo, Suk Hyun [Dept. of Radiation Oncology, Seoul National University Hospital, Seoul (Korea, Republic of); Seong, Won Mo [Interdisciplinary Program of Radiation Applied Life Science, Seoul National University College of Medicine, Seoul (Korea, Republic of); Kim, Kyu Bo [Dept. of Radiation Oncology, Seoul National University College of Medicine, Seoul (Korea, Republic of)

    2012-09-15

    In this study, we considerate our radiation therapy process for the breast cancer patient implanted a pacemaker applying the machine movement surgery, shielding, beam selection. We perform radiation therapy to a 54 years old, breast cancer patient implanted a pacemaker. The patient underwent a surgery to move the position of a pacemaker to right side breast after consultation with cardiology department. Prescribed dose was 5,040 cGy and daily dose 180 cGy for 28 fractions. The 10 MV photon energy, field size 0/9.5x20 cm, half beam and opposing portal irradiation are used. To find out appropriate thickness of shielding board, we carried out an experiment using a solid water phantom (30x30x7 cm), a Farmer-type chamber (TN30013, PTW, Germany) and a shielding board (Pb 28x27x0.1 cm). We calculated expected absorbed dose to te pacemaker with absorb ratio and shielding ratio. In the PTP system (Eclipse, Varian, USA), we figured out how much radiation would be absorbed to the machine with and without shielding. First day of the radiation therapy, we measured head scatter to the pacemaker with MOSFET Dose Verification System (TN-RD-70-W, Medical Canada Ltd., Canada). In the phantom measurement, we found out appropriate thickness was 2 mm of shielding board. In the RTP, when using 2 mm shielding the pacemaker will be absorbed 11.5-38.2 cGy and DVH is 77.3 cGy. In the first day of the therapy, 4.3 cGy was measured so 120.4 cGy was calculated during total therapy. The patient was free from any side effects, and the machine also normally functioned. As the report of association which have public confidence became superannuated, there is lack of data about new machine. We believe that radiation therapy to these kind of patients could be done successfully with co-operation, patient-suitable planning, accurate QA, frequent in-vivo dosimetry and monitoring.

  4. Cardiac Pacemakers; Marcapasos Cardiacos

    Energy Technology Data Exchange (ETDEWEB)

    Fiandra, O [Universidad de la Republica, Facultad de Maedicina, Departamento de Cardiologia, Montevideo(Uruguay); Espasandin, W [Universidad de la Republica, Facultad de Medicina, Departamento de Cirugia Cardiaca, Montevideo (Uruguay); Fiandra, H [Instituto Nacional de Cirugia Cardiaca, Departamento de Hemodinamia y Marcapasos, Montevideo (Uruguay); and others

    1984-07-01

    A complete survey of physiological biophysical,clinical and engineering aspects of cardiac facing,including the history and an assessment of possible future developments.Among the topics studied are: pacemakers, energy search, heart stimulating with pacemakers ,mathematical aspects of the electric cardio stimulation chronic, pacemaker implants,proceeding,treatment and control.

  5. Evaluation of the effect of different types and location of pacemaker on cardiac function using radionuclide techniques

    International Nuclear Information System (INIS)

    Shi Rongfang; Liu Xiujie; Guo Xiuzhen

    1993-01-01

    Left ventricular function was measured in 24 patients using radionuclide ventriculography before and after planting different types of pacemakers. The results showed that the heart rate, peak filling rate (PFR) and relative cardiac output (RCO) were significantly increased after planting the pacemaker. However, the effects of AAI and VVI on the synchronism of LV were different. AAI can significantly improve the synchronism and compliance of LV

  6. Emergency Management of Malignant Acute Left-Sided Colonic Obstruction

    Science.gov (United States)

    Trompetas, Vasileios

    2008-01-01

    INTRODUCTION The management of acute left-sided colonic obstruction still remains a challenging problem despite significant progress. METHODS A literature search was undertaken using PubMed and the Cochrane Library regarding the options in emergency management of left-sided colonic obstruction focusing on outcomes such as mortality, morbidity, long-term prognosis and cost effectiveness. DISCUSSION Colonic stenting is the best option either for palliation or as a bridge to surgery. It reduces morbidity and mortality rate and the need for colostomy formation. Stenting is likely to be cost effective, but data are variable depending on the individual healthcare system. Nevertheless, surgical management remains relevant as colonic stenting has a small rate of failure, and it is not always available. There are various surgical options. One-stage primary resection and anastomosis is the preferred choice for low-risk patients. Intra-operative colonic irrigation has no proven benefit. Subtotal colectomy is useful in cases of proximal bowel damage or synchronous tumours. Hartmann's procedure should be reserved for high-risk patients. Simple colostomy has no role other than for use in very ill patients who are not fit for any other procedure. PMID:18430330

  7. Left-sided incarcerated Amyand’s hernia with cecum and terminal ileum: a case report

    Directory of Open Access Journals (Sweden)

    Bekele K

    2017-10-01

    Full Text Available Kebebe Bekele,1 Desalegn Markos2 1Department of Surgery, School of Medicine, Madda Walabu University, Bale Robe, Ethiopia; 2Unit of Neonatology Nursing, St Paul Hospital Millennium Medical College, Addis Ababa, Ethiopia Background: Amyand’s hernia, which is the presence of a normal or pathological appendix as a part of an inguinal hernia, is a rare clinical entity. We are reporting a very rare case of left-sided incarcerated Amyand’s hernia with cecum and terminal ileum involvement. Case presentation: A 4-year-old male child with left inguinal swelling of 2-year duration presented to Goba Referral Hospital. Two days before the patient visited our hospital, the swelling had become irreducible and caused severe groin pain. He had abdominal cramps, bilious vomiting, and mild abdominal distention, but passed feces. With the diagnosis of left-sided incarcerated inguinal hernia, the patient was investigated and prepared for surgical management. During the operative procedure, we identified the presence of appendix, cecum, and terminal ileum in the scrotum as the herniated component. After the sack was dissected, since there was also appendicitis, an appendectomy was performed. Then, high ligation of sack was done after cecum and ileum were reduced. After 3 uneventful postoperative days in the hospital, the patient was discharged. The patient was followed-up for 6 months, and he did not develop any complications. Conclusion: Left-sided incarcerated Amyand’s hernia with cecum and terminal ileum involvement is a rare clinical entity. Even though it is not common, appendicitis is one of the comorbidities that can be seen in patients with left-sided incarcerated Amyand’s hernia with cecum and terminal ileum. Surgeons should have a high index of clinical suspicion and be aware of the potential involvement of appendix, cecum, and ileum as part of an incarcerated hernia during surgery, even in the left inguinal region. In this case, left-sided

  8. Left sided circumaortic and retroaortic left renal veins, renal artery arising from iliac common artery in L-shaped kidney

    International Nuclear Information System (INIS)

    Al-Amin, M.

    2012-01-01

    Full text: Introduction: Renal ectopia is a congenital anomaly with variable clinical presentation. Kidneys are normally located in the retroperitoneal position, on either side of vertebral column, against the psoas muscles but when not at such position, it is called renal ectopia or ectopic kidney. Ectopic kidneys are thought to occur in approximately 1 in 1,000 births but only about 1 in 10 of these is ever diagnosed. In 90% of crossed ectopy, there is at least partial fusion of the kidneys. Left-to right ectopy is thought to be three times more common. Some of these are discovered incidentally, when a child or adult is having ultrasonography for a medical condition unrelated to renal ectopia. In a crossed fused renal ectopic kidney, complications such as nephrolithiasis, infection, and hydronephrosis approaches over 50%. Simple renal ectopia refers to kidney that is located on the proper side but abnormal in position. Crossed renal ectopia was first described by Pannorlus in 1964 and refer to kidney that has crossed from left to right or vice-versa, with moving of one kidney to the opposite side following ascent of the other kidney, so that both kidneys are located on the same side of the body, mostly fused called crossed fused ectopia. The fusion of the two kidneys is believed to result from (1) failure of the primitive nephrogenic cell masses to separate or (2) fusion of the two blastemas during their abdominal ascent. Discussion: A 57-year-old woman with a new found hematological disease. CT exam was performed with intravenous application of contrast media. Like an additional findings we visualized the presence of right to-left ectopy (L - shaped kidney) and the presence of left circumaortic renal vein emanating from a normally situated left kidney and retroaortic renal vein as having been located by the ectopic right kidney. Conclusion: By crossed renal ectopia is meant congenital displacement of one kidney to the opposite side. The conditional may present

  9. Programmable Pacemaker

    Science.gov (United States)

    1996-01-01

    Released in 1995, the Trilogy cardiac pacemaker is the fourth generation of a unit developed in the 1970s by NASA, Johns Hopkins Applied Physics Laboratory and St. Jude Medical's Cardiac Rhythm Management Division (formerly known as Pacesetter Systems, Inc.). The new system incorporates the company's PDx diagnostic and programming software and a powerful microprocessor that allows more functions to be fully automatic and gives more detailed information on the patient's health and the performance of the pacing systems. The pacemaker incorporates bidirectional telemetry used for space communications for noninvasive communication with the implanted pacemaker, smaller implantable pulse generators from space microminiaturization, and longer-life batteries from technology for spacecraft electrical power systems.

  10. Left-sided congenital heart lesions in mosaic Turner syndrome.

    Science.gov (United States)

    Bouayed Abdelmoula, Nouha; Abdelmoula, Balkiss; Smaoui, Walid; Trabelsi, Imen; Louati, Rim; Aloulou, Samir; Aloulou, Wafa; Abid, Fatma; Kammoun, Senda; Trigui, Khaled; Bedoui, Olfa; Denguir, Hichem; Mallek, Souad; Ben Aziza, Mustapha; Dammak, Jamila; Kaabi, Oldez; Abdellaoui, Nawel; Turki, Fatma; Kaabi, Asma; Kamoun, Wafa; Jabeur, Jihen; Ltaif, Wided; Chaker, Kays; Fourati, Haytham; M'rabet, Samir; Ben Ameur, Hedi; Gouia, Naourez; Mhiri, Mohamed Nabil; Rebai, Tarek

    2018-04-01

    In the era of the diseasomes and interactome networks, linking genetics with phenotypic traits in Turner syndrome should be studied thoroughly. As a part of this stratagem, mosaicism of both X and Y chromosome which is a common finding in TS and an evaluation of congenital heart diseases in the different situations of mosaic TS types, can be helpful in the identification of disturbed sex chromosomes, genes and signaling pathway actors. Here we report the case of a mosaic TS associated to four left-sided CHD, including BAV, COA, aortic aneurysms and dissections at an early age. The mosaicism included two cell lines, well-defined at the cytogenetic and molecular levels: a cell line which is monosomic for Xp and Xq genes (45,X) and another which is trisomic for pseudoautosomal genes that are present on the X and Y chromosomes and escape X inactivation: 45,X[8]/46,X,idic(Y)(pter→q11.2::q11.2→pter)[42]. This case generates two hypotheses about the contribution of genes linked to the sex chromosomes and the signaling pathways involving these genes, in left-sided heart diseases. The first hypothesis suggests the interaction between X chromosome and autosomal genes or loci of aortic development, possibly dose-dependent, and which could be in the framework of TGF-β-SMAD signaling pathways. The second implies that left-sided congenital heart lesions involve sex chromosomes loci. The reduced dosage of X chromosome gene(s), escaping X inactivation during development, contributes to this type of CHD. Regarding our case, these X chromosome genes may have homologues at the Y chromosome, but the process of inactivation of the centromeres of the isodicentric Y spreads to the concerned Y chromosome genes. Therefore, this case emerges as an invitation to consider the mosaics of Turner syndrome and to study their phenotypes in correlation with their genotypes to discover the underlying developmental and genetic mechanisms, especially the ones related to sex chromosomes.

  11. Pre-ejection period by radial artery tonometry supplements echo doppler findings during biventricular pacemaker optimization

    Directory of Open Access Journals (Sweden)

    Qamruddin Salima

    2011-07-01

    Full Text Available Abstract Background Biventricular (Biv pacemaker echo optimization has been shown to improve cardiac output however is not routinely used due to its complexity. We investigated the role of a simple method involving computerized pre-ejection time (PEP assessment by radial artery tonometry in guiding Biv pacemaker optimization. Methods Blinded echo and radial artery tonometry were performed simultaneously in 37 patients, age 69.1 ± 12.8 years, left ventricular (LV ejection fraction (EF 33 ± 10%, during Biv pacemaker optimization. Effect of optimization on echo derived velocity time integral (VTI, ejection time (ET, myocardial performance index (MPI, radial artery tonometry derived PEP and echo-radial artery tonometry derived PEP/VTI and PEP/ET indices was evaluated. Results Significant improvement post optimization was achieved in LV ET (286.9 ± 37.3 to 299 ± 34.6 ms, p Conclusion An acute shortening of PEP by radial artery tonometry occurs post Biv pacemaker optimization and correlates with improvement in hemodynamics by echo Doppler and may provide a cost-efficient approach to assist with Biv pacemaker echo optimization.

  12. Clinical application of transvenous temporary cardiac pacemaker in performing extra-cranial carotid angiography and stent implantation

    International Nuclear Information System (INIS)

    Liu Juan; Yao Guoen; Zhou Huadong; Jiang Xiaojiang; Chen Qiao

    2012-01-01

    Objective: To assess the safety and effectiveness of transvenous temporary cardiac pacemaker in preventing hemodynamic instability occurred during the perioperative period of extra-cranial carotid angiography and stent implantation. Methods: Preoperative install of temporary cardiac pacemaker via left femoral vein was carried out in 41 patients who were at high risk for developing hemodynamic instability, which was followed by extra-cranial carotid angiography and stent implantation. The pacing rhythm of the pacemaker was fixed at 60 beats/min. During and after the procedure the patients were under close observation for the signs of discomfort symptoms as well as the changes in blood pressure and heart rate. The working condition of the pacemaker was also monitored. Results: All the installed pacemakers were technically and hemodynamically effective in producing electrical ventricular responses in all 25 patients who had received balloon dilatation of carotid in advance. Transient pacemaker activation appeared in 25 patients. The longest activation time was one day. During pacemaker activation, one patient developed symptomatic hypotension. The longest duration of hypotension lasted for 4 days. No pacemaker-related or procedure-related complications occurred. Conclusion: Hemodynamic instability is a common complication occurred during perioperative period of extra-cranial carotid angiography and stent implantation. As a prophylactic measure, preoperative placement of temporary cardiac pacemaker can promptly and effectively correct the hemodynamic disorders and prevent perioperative complications such as stroke, etc. Therefore, this technique is worth employing in clinical practice, and it is especially useful for patients with high risks. (authors)

  13. Anterior subcarinal node dissection on the left side using video thoracoscopy: an easier technique.

    Science.gov (United States)

    Baste, Jean-Marc; Haddad, Laura; Melki, Jean; Peillon, Christophe

    2015-04-01

    Lobectomy for lung carcinoma is usually associated with complete node dissection, but it is often difficult to perform using video thoracoscopy, especially on the left side. In this case, our team uses an anterior technique for subcarinal lymphadenectomy. After left lobectomy, we lift the bronchial stump by its anterior face to open and dissect the subcarinal space. Exposure is difficult using the more usual technique of posterior subcarinal lymphadenectomy, and the different techniques (often requiring retractors) remain complex because some vessels might be injured. We recommend using anterior lymphadenectomy, which should facilitate video thoracoscopy for lymphadenectomy on the left side. Copyright © 2015 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  14. [Dextrals and sinistrals (right-handers and left-handers): specificity of interhemispheric brain asymmetry and EEG coherence parameters].

    Science.gov (United States)

    Zhavoronkova, L A

    2007-01-01

    Data of literature about morphological, functional and biochemical specificity of the brain interhemispheric asymmetry of healthy right-handers and left-handers and about peculiarity of dynamics of cerebral pathology in patients with different individual asymmetry profiles are presented at the present article. Results of our investigation by using coherence parameters of electroencephalogram (EEG) in healthy right-handers and left-handers in state of rest, during functional tests and sleeping and in patients with different forms of the brain organic damage were analyzed too. EEG coherence analysis revealed the reciprocal changing of alpha-beta and theta-delta spectral bands in right-handers whilein left-handers synchronous changing of all EEG spectral bands were observed. Data about regional-frequent specificity of EEG coherence, peculiarity of EEG asymmetry in right-handers and left-handers, aslo about specificity of EEG spectral band genesis and point of view about a role of the brain regulator systems in forming of interhemispheric asymmetry in different functional states allowed to propose the conception about principle of interhermispheric brain asymmetry formation in left-handers and left-handers. Following this conception in dextrals elements of concurrent (summary-reciprocal) cooperation are predominant at the character of interhemispheric and cortical-subcortical interaction while in sinistrals a principle of concordance (supplementary) is preferable. These peculiarities the brain organization determine, from the first side, the quicker revovery of functions damaged after cranio-cerebral trauma in left-handers in comparison right-handers and from the other side - they determine the forming of the more expressed pathology in the remote terms after exposure the low dose of radiation.

  15. Incidence and factors associated with infective endocarditis in patients undergoing left-sided heart valve replacement

    DEFF Research Database (Denmark)

    Østergaard, Lauge; Valeur, Nana; Ihlemann, Nikolaj

    2018-01-01

    Aims: Patients with left-sided heart valve replacement are considered at high-risk of infective endocarditis (IE). However, data on the incidence and risk factors associated with IE are sparse. Methods and results: Through Danish administrative registries, we identified patients who underwent left.......35-2.15), and cardiac implantable electronic device (CIED) (HR = 1.57, 95% CI 1.19-2.06) were among factors associated with an increased risk of IE. Conclusion: Infective endocarditis after left-sided heart valve replacement is not uncommon and occurs in about 1/20 over 10 years. Male, bioprosthetic valve, and heart...

  16. Right colon cancer: Left behind.

    Science.gov (United States)

    Gervaz, P; Usel, M; Rapiti, E; Chappuis, P; Neyroud-Kaspar, I; Bouchardy, C

    2016-09-01

    Prognosis of colon cancer (CC) has steadily improved during the past three decades. This trend, however, may vary according to proximal (right) or distal (left) tumor location. We studied if improvement in survival was greater for left than for right CC. We included all CC recorded at the Geneva population-based registry between 1980 and 2006. We compared patients, tumor and treatment characteristics between left and right CC by logistic regression and compared CC specific survival by Cox models taking into account putative confounders. We also compared changes in survival between CC location in early and late years of observation. Among the 3396 CC patients, 1334 (39%) had right-sided and 2062 (61%) left-sided tumors. In the early 1980s, 5-year specific survival was identical for right and left CCs (49% vs. 48%). During the study period, a dramatic improvement in survival was observed for patients with left-sided cancers (Hazard ratio [HR]: 0.42, 95% confidence interval [CI]: 0.29-0.62, p colon cancer patients, those with right-sided lesions have by far the worse prognosis. Change of strategic management in this subgroup is warranted. Copyright © 2016 Elsevier Ltd. All rights reserved.

  17. Pacemaker Use Following Heart Transplantation

    Science.gov (United States)

    Mallidi, Hari R.; Bates, Michael

    2017-01-01

    Background: The incidence of permanent pacemaker implantation after orthotopic heart transplantation has been reported to be 2%-24%. Transplanted hearts usually exhibit sinus rhythm in the operating room following reperfusion, and most patients do not exhibit significant arrhythmias during the postoperative period. However, among the patients who do exhibit abnormalities, pacemakers may be implanted for early sinus node dysfunction but are rarely used after 6 months. Permanent pacing is often required for atrioventricular block. A different cohort of transplant patients presents later with bradycardia requiring pacemaker implantation, reported to occur in approximately 1.5% of patients. The objectives of this study were to investigate the indications for pacemaker implantation, compare the need for pacemakers following bicaval vs biatrial anastomosis, and examine the long-term outcomes of heart transplant patients who received pacemakers. Methods: For this retrospective, case-cohort, single-institution study, patients were identified from clinical research and administrative transplant databases. Information was supplemented with review of the medical records. Standard statistical techniques were used, with chi-square testing for categorical variables and the 2-tailed t test for continuous variables. Survival was compared with the use of log-rank methods. Results: Between January 1968 and February 2008, 1,450 heart transplants were performed at Stanford University. Eighty-four patients (5.8%) were identified as having had a pacemaker implanted. Of these patients, 65.5% (55) had the device implanted within 30 days of transplantation, and 34.5% (29) had late implantation. The mean survival of patients who had an early pacemaker implant was 6.4 years compared to 7.7 years for those with a late pacemaker implant (Ppacemaker implantation. Starting in 1997, a bicaval technique was used for implantation. The incidence of pacemaker implantation by technique was 2.0% for

  18. Transcriptional analysis of left-sided colitis, pancolitis, and ulcerative colitis-associated dysplasia

    DEFF Research Database (Denmark)

    Bjerrum, Jacob T; Nielsen, Ole H; Riis, Lene B

    2014-01-01

    to identify potential biomarkers and transcripts of importance for the carcinogenic behavior of chronic inflammation. METHODS: The Affymetrix GeneChip Human Genome U133 Plus 2.0 was applied on colonic biopsies from UC patients with left-sided UC, pancolitis, dysplasia, and controls. Reverse transcription...... polymerase chain reaction and immunohistochemistry were performed for validating selected transcripts in the initial cohort and in 2 independent cohorts of patients with UC. Microarray data were analyzed by principal component analysis, and reverse transcription polymerase chain reaction...... and immunohistochemistry data by the Wilcoxon's rank-sum test. RESULTS: The principal component analysis results revealed separate clusters for left-sided UC, pancolitis, dysplasia, and controls. Close clustering of dysplastic and pancolitic samples indicated similarities in gene expression. Indeed, 101 and 656 parallel...

  19. Is Diversion with Ileostomy Non-inferior to Hartmann Resection for Left-sided Colorectal Anastomotic Leak?

    Science.gov (United States)

    Stafford, Caitlin; Francone, Todd D; Marcello, Peter W; Roberts, Patricia L; Ricciardi, Rocco

    2018-03-01

    Treatment of left-sided colorectal anastomotic leaks often requires fecal stream diversion for prevention of further septic complications. To manage anastomotic leak, it is unclear if diverting ileostomy provides similar outcomes to Hartmann resection with colostomy. We identified all patients who developed anastomotic leak following left-sided colorectal resections from 1/2012 through 12/2014 using the American College of Surgeons National Surgical Quality Improvement Program. Then, we examined the risk of mortality and abdominal reoperation in patients treated with diverting ileostomy as compared to Hartmann resection. There were 1745 patients who experienced an anastomotic leak in a cohort of 63,748 patients (3.7%). Two hundred thirty-five patients had a reoperation for anastomotic leak involving the formation of a diverting ileostomy (n = 77) or Hartmann resection (n = 158). There was no difference in mortality or abdominal reoperation in patients treated with diverting ileostomy (3.9, 7.8%) versus Hartmann resection (3.8, 6.3%) (p = 0.8). There was no difference in the outcomes of mortality or need for second abdominal reoperation in patients treated with diverting ileostomy as compared to Hartmann resection for left-sided colorectal anastomotic leak. Thus, select patients with left-sided colorectal anastomotic leaks may be safely managed with diverting ileostomy.

  20. Guidelines of diagnostics and treatment of acute left-sided colonic diverticulitis

    NARCIS (Netherlands)

    Andeweg, Caroline S.; Mulder, Irene M.; Felt-Bersma, Richelle J. F.; Verbon, Annelies; van der Wilt, Gert Jan; van Goor, Harry; Lange, Johan F.; Stoker, Jaap; Boermeester, Marja A.; Bleichrodt, Robert P.

    2013-01-01

    The incidence of acute left-sided colonic diverticulitis (ACD) is increasing in the Western world. To improve the quality of patient care, a guideline for diagnosis and treatment of diverticulitis is needed. A multidisciplinary working group, representing experts of relevant specialties, was

  1. A Left-Sided Prevalence of Lentigo Maligna: A UK Based Observational Study and Review of the Evidence

    Directory of Open Access Journals (Sweden)

    Mark Gorman

    2015-01-01

    Full Text Available Skin cancer has been shown to present asymmetrically, prevalent on the left side of the body, more so in subtypes of cutaneous melanoma such as lentigo maligna. Biases have been linked to cumulative UV light exposure and automobile driving patterns. Though left-right ratios have previously correlated with the side men or women tend to position themselves or countries drive on, more recent trends indicate a consistent left-sided bias. To clarify reasons for changing trends, a review of the evidence base and LM’s laterality in a UK cohort (99 cases 2000–2011 was conducted for the first time. The strong correlation of left-sided excess, found in both genders (ratios 1.381–1.5, P<0.05  X2 0.841, is congruent with more recent findings. Though evidence indicates that driving position is no longer a risk factor for LM, due most likely to improved car window UV protection, it remains the most commonly attributed cause. Understanding phenomena such as UV lights “scatter effect” or that cumulative exposure may not be a significant risk factor helps rationalize older conclusions that would otherwise appear contradictory. The reasons for left-sided excess remain unclear but may be due to factors requiring further research such as the body’s anatomical/embryological asymmetry.

  2. Trends in Cardiac Pacemaker Batteries

    Directory of Open Access Journals (Sweden)

    Venkateswara Sarma Mallela

    2004-10-01

    Full Text Available Batteries used in Implantable cardiac pacemakers-present unique challenges to their developers and manufacturers in terms of high levels of safety and reliability. In addition, the batteries must have longevity to avoid frequent replacements. Technological advances in leads/electrodes have reduced energy requirements by two orders of magnitude. Micro-electronics advances sharply reduce internal current drain concurrently decreasing size and increasing functionality, reliability, and longevity. It is reported that about 600,000 pacemakers are implanted each year worldwide and the total number of people with various types of implanted pacemaker has already crossed 3 million. A cardiac pacemaker uses half of its battery power for cardiac stimulation and the other half for housekeeping tasks such as monitoring and data logging. The first implanted cardiac pacemaker used nickel-cadmium rechargeable battery, later on zinc-mercury battery was developed and used which lasted for over 2 years. Lithium iodine battery invented and used by Wilson Greatbatch and his team in 1972 made the real impact to implantable cardiac pacemakers. This battery lasts for about 10 years and even today is the power source for many manufacturers of cardiac pacemakers. This paper briefly reviews various developments of battery technologies since the inception of cardiac pacemaker and presents the alternative to lithium iodine battery for the near future.

  3. Cardiovascular malformations caused by NOTCH1 mutations do not keep left : data on 428 probands with left-sided CHD and their families

    NARCIS (Netherlands)

    Kerstjens-Frederikse, Wilhelmina S.; van de Laar, Ingrid M. B. H.; Vos, Yvonne J.; Verhagen, Judith M. A.; Berger, Rolf M. F.; Lichtenbelt, Klaske D.; Wassink-Ruiter, Jolien S. Klein; van der Zwaag, Paul A.; Sarvaas, Gideon J. du Marchie; Bergman, Klasien A.; Bilardo, Catia M.; Roos-Hesselink, Jolien W.; Janssen, Johan H. P.; Frohn-Mulder, Ingrid M.; van Spaendonck-Zwarts, Karin Y.; Melle, van Joost P.; Hofstra, Robert M. W.; Wessels, M. W.

    Purpose: We aimed to determine the prevalence and phenotypic spectrum of NOTCH1 mutations in left-sided congenital heart disease (LS-CHD). LS-CHD includes aortic valve stenosis, a bicuspid aortic valve, coarctation of the aorta, and hypoplastic left heart syndrome. Methods: NOTCH1 was screened for

  4. Calcitriol Supplementation Causes Decreases in Tumorigenic Proteins and Different Proteomic and Metabolomic Signatures in Right versus Left-Sided Colon Cancer

    Directory of Open Access Journals (Sweden)

    Monica M. Schroll

    2018-01-01

    Full Text Available Vitamin D deficiency is a common problem worldwide. In particular, it is an issue in the Northern Hemisphere where UVB radiation does not penetrate the atmosphere as readily. There is a correlation between vitamin D deficiency and colorectal cancer incidence and mortality. Furthermore, there is strong evidence that cancer of the ascending (right side colon is different from cancer of the descending (left side colon in terms of prognosis, tumor differentiation, and polyp type, as well as at the molecular level. Right-side tumors have elevated Wnt signaling and are more likely to relapse, whereas left-side tumors have reduced expression of tumor suppressor genes. This study seeks to understand both the proteomic and metabolomic changes resulting from treatment of the active metabolite of vitamin D, calcitriol, in right-sided and left-sided colon cancer. Our results show that left-sided colon cancer treated with calcitriol has a substantially greater number of changes in both the proteome and the metabolome than right-sided colon cancer. We found that calcitriol treatment in both right-sided and left-sided colon cancer causes a downregulation of ribosomal protein L37 and protein S100A10. Both of these proteins are heavily involved in tumorigenesis, suggesting a possible mechanism for the correlation between low vitamin D levels and colon cancer.

  5. Calcitriol Supplementation Causes Decreases in Tumorigenic Proteins and Different Proteomic and Metabolomic Signatures in Right versus Left-Sided Colon Cancer.

    Science.gov (United States)

    Schroll, Monica M; Ludwig, Katelyn R; Bauer, Kerry M; Hummon, Amanda B

    2018-01-11

    Vitamin D deficiency is a common problem worldwide. In particular, it is an issue in the Northern Hemisphere where UVB radiation does not penetrate the atmosphere as readily. There is a correlation between vitamin D deficiency and colorectal cancer incidence and mortality. Furthermore, there is strong evidence that cancer of the ascending (right side) colon is different from cancer of the descending (left side) colon in terms of prognosis, tumor differentiation, and polyp type, as well as at the molecular level. Right-side tumors have elevated Wnt signaling and are more likely to relapse, whereas left-side tumors have reduced expression of tumor suppressor genes. This study seeks to understand both the proteomic and metabolomic changes resulting from treatment of the active metabolite of vitamin D, calcitriol, in right-sided and left-sided colon cancer. Our results show that left-sided colon cancer treated with calcitriol has a substantially greater number of changes in both the proteome and the metabolome than right-sided colon cancer. We found that calcitriol treatment in both right-sided and left-sided colon cancer causes a downregulation of ribosomal protein L37 and protein S100A10. Both of these proteins are heavily involved in tumorigenesis, suggesting a possible mechanism for the correlation between low vitamin D levels and colon cancer.

  6. Value of local electrogram characteristics predicting successful catheter ablation of left-versus right-sided accessory atrioventricular pathways by radiofrequency current.

    Science.gov (United States)

    Lin, J L; Schie, J T; Tseng, C D; Chen, W J; Cheng, T F; Tsou, S S; Chen, J J; Tseng, Y Z; Lien, W P

    1995-01-01

    Despite similar guidance by local electrogram criteria, catheter ablation of right-sided accessory atrioventricular (AV) pathways by radiofrequency current has been less effective than that of left-sided ones. In order to elucidate the possible diversities in local electrosignal criteria, we systematically analyzed the morphological and timing characteristics of 215 bipolar local electrograms from catheter ablation sites of 65 left-sided accessory AV pathways and of 356 from those of 37 right-sided ones in 92 consecutive patients with Wolff-Parkinson-White syndrome or AV reentrant tachycardia incorporating concealed accessory AV pathways. After stepwise multivariate analysis, we selected the presence of a possible accessory pathway potential, local ventricular activation preceding QRS complex for 20 ms or more during ventricular insertion mapping, and the local retrograde ventriculoatrial (VA) continuity, local retrograde VA interval right-sided targets only) during atrial insertion mapping, as independent local electrogram predictors for successful ablation of left- and right-sided accessory AV pathways. Combination of all local electrogram predictors could have moderate chance of success (80 and 51%) for the ventricular and atrial insertion ablation of left-sided accessory AV pathways, but only low probability of success (40% in ventricular insertion ablation) or very low sensitivity (12.5% in atrial insertion ablation) for right-sided ones. In conclusion, with the present approach, successful catheter ablation of right-sided accessory AV pathways, compared to left-sided ones, still necessitate a breakthrough in the precision mapping and the efficiency of energy delivery.

  7. Radiation effect on implanted pacemakers

    International Nuclear Information System (INIS)

    Pourhamidi, A.H.

    1983-01-01

    It was previously thought that diagnostic or therapeutic ionizing radiation did not have an adverse effect on the function of cardiac pacemakers. Recently, however, some authors have reported damaging effect of therapeutic radiation on cardiac pulse generators. An analysis of a recently-extracted pacemaker documented the effect of radiation on the pacemaker pulse generator

  8. Does Side Make a Difference? Anatomical Differences Between the Left and Right Ureter.

    Science.gov (United States)

    Odegard, Stephen E; Abernethy, Melinda G; Mueller, Elizabeth R

    2015-01-01

    Seventy to eighty percent of iatrogenic ureteral injuries involve the left ureter. We sought to evaluate potential anatomical differences between the left and right ureters that may contribute to this discrepancy. A retrospective image review was undertaken of women who underwent computed tomography urograms between 2012 and 2013. The distance to the ureters from the midline was measured at the level of the sacral promontory (S1) and the cervix. Cervical deviation from the midline was measured, and distance between the cervix and ureters was calculated. The anterior-posterior distance between ureters was also measured. Ninety-five computed tomography urograms were analyzed. The mean age was 56 years (range, 23-92 years). Mean cervical deviation was 2.9 mm left of the midline (P = 0.028). The left ureter was 4.2 mm more lateral than the right at S1 and 2.7 mm more lateral at the cervix (P = 0.000 and 0.001). There was no significant difference when accounting for cervical deviation (P = 0.220). The left ureter was 1.9 mm more anterior than the right at the cervix (P = 0.012). Age, body mass index, and ethnicity did not affect the ureteral position. Based on midline measurements, the left ureter courses 2 to 4 mm more lateral and anterior than does the right ureter. The cervix is also positioned 2 to 4 mm to the left side, and as a result, the ureters are actually symmetric to the cervix. Although seemingly small, 2 to 4 mm is the width range of a Heaney clamp. These anatomic differences may be a contributing factor to the increase in ureteral injuries on the left side compared with the right.

  9. Ventricular fibrillation induced by coagulating mode bipolar electrocautery during pacemaker implantation in Myotonic Dystrophy type 1 patient.

    Science.gov (United States)

    Russo, Vincenzo; Rago, Anna; DI Meo, Federica; Cioppa, Nadia Della; Papa, Andrea Antonio; Russo, Maria Giovanna; Nigro, Gerardo

    2014-12-01

    The occurrence of ventricular fibrillation, induced by bipolar electrocautery during elective dual chamber pacemaker implantation, is reported in a patient affected by Myotonic Distrophy type 1 with normal left ventricular ejection fraction.

  10. A case of hypoplasia of left lung with very rare associations with congenital absence of left pulmonary artery and right-sided aortic arch

    Directory of Open Access Journals (Sweden)

    Trilok Chand

    2017-01-01

    Full Text Available The absence of one of the pulmonary artery with associated hypoplasia of lung and great vessel abnormality is a rare finding. The incidence of this rare congenital abnormality is around 1 in 200,000 live birth. The absence of the left side pulmonary artery is again uncommon, and associated cardiac malformations are usually tetralogy of fallot or septal defects rather than an aortic arch defect. Our case is a unique case in It’s associated congenital anomalies. He was presented with recurrent pneumothorax and hemoptysis, and on thorough workup, he was diagnosed to have an absence of left pulmonary artery with hypoplasia of the left lung and associated right-sided aortic arch. The patient’s family has declined the surgical option, and he was managed conservatively and kept in close follow-up.

  11. [Left-sided endocarditis due to gram-negative bacilli: epidemiology and clinical characteristics].

    Science.gov (United States)

    Noureddine, Mariam; de la Torre, Javier; Ivanova, Radka; Martínez, Francisco José; Lomas, Jose María; Plata, Antonio; Gálvez, Juan; Reguera, Jose María; Ruiz, Josefa; Hidalgo, Carmen; Luque, Rafael; García-López, María Victoria; de Alarcón, Arístides

    2011-04-01

    The aim of this study is to describe the epidemiological, clinical characteristics, and outcome of patients with left-side endocarditis caused by gram-negative bacteria. Prospective multicenter study of left-sided infective endocarditis reported in the Andalusian Cohort for the Study of Cardiovascular Infections between 1984 and 2008. Among the 961 endocarditis, 24 (2.5%) were caused by gram-negative bacilli. The most common pathogens were Escherichia coli, Pseudomonas aeruginosa and Salmonella enterica. Native valves (85.7%) were mainly affected, most of them with previous valve damage (57%). Comorbidity was greater (90% vs 39%; P=.05) than in endocarditis due to other microorganism, the most frequent being, diabetes, hepatic cirrhosis and neoplasm. A previous manipulation was found in 47.6% of the cases, and 37% were considered hospital-acquired. Renal failure (41%), central nervous system involvement (33%) and ventricular dysfunction (45%) were the most frequent complications. Five cases (21%) required cardiac surgery, mostly due to ventricular dysfunction. More than 50% of cases were treated with aminoglycosides, but this did not lead to a better outcome or prognosis. Mortality (10 patients) was higher than that reported with other microorganisms (41% vs 35%; P=.05). Left-sided endocarditis due to gram-negative bacilli is a rare disease, which affects patients with major morbidities and often with a previous history of hospital manipulations. Cardiac, neurological and renal complications are frequent and associated with a high mortality. The association of aminoglycosides in the antimicrobial treatment did not involve a better outcome or prognosis. Copyright © 2010 Elsevier España, S.L. All rights reserved.

  12. Damaging effect of therapeutic radiation on programmable pacemakers

    International Nuclear Information System (INIS)

    Adamec, R.; Haefliger, J.M.; Killisch, J.P.; Niederer, J.; Jaquet, P.

    1982-01-01

    Two series of present-day pacemakers were tested in vitro with pulsed x-ray radiation. The first series of 12 pacemakers consisted of 10 different types and models of demand pacemakers (VVI). The second series of 13 pacemakers had 9 different types and models of programmable pacemakers. Unlike the first series which showed only mild changes in frequency and pulse width, all but four of the programmable pacemakers presented sudden complete failure after different radiation doses. We conclude that direct pulse radiation at therapeutic levels of programmable pacemakers should be avoided

  13. Inward-rectifying potassium (Kir) channels regulate pacemaker activity in spinal nociceptive circuits during early life

    Science.gov (United States)

    Li, Jie; Blankenship, Meredith L.; Baccei, Mark L.

    2013-01-01

    Pacemaker neurons in neonatal spinal nociceptive circuits generate intrinsic burst-firing and are distinguished by a lower “leak” membrane conductance compared to adjacent, non-bursting neurons. However, little is known about which subtypes of leak channels regulate the level of pacemaker activity within the developing rat superficial dorsal horn (SDH). Here we demonstrate that a hallmark feature of lamina I pacemaker neurons is a reduced conductance through inward-rectifying potassium (Kir) channels at physiological membrane potentials. Differences in the strength of inward rectification between pacemakers and non-pacemakers indicate the presence of functionally distinct Kir currents in these two populations at room temperature. However, Kir currents in both groups showed high sensitivity to block by extracellular Ba2+ (IC50 ~ 10 µM), which suggests the presence of ‘classical’ Kir (Kir2.x) channels in the neonatal SDH. The reduced Kir conductance within pacemakers is unlikely to be explained by an absence of particular Kir2.x isoforms, as immunohistochemical analysis revealed the expression of Kir2.1, Kir2.2 and Kir2.3 within spontaneously bursting neurons. Importantly, Ba2+ application unmasked rhythmic burst-firing in ~42% of non-bursting lamina I neurons, suggesting that pacemaker activity is a latent property of a sizeable population of SDH cells during early life. In addition, the prevalence of spontaneous burst-firing within lamina I was enhanced in the presence of high internal concentrations of free Mg2+, consistent with its documented ability to block Kir channels from the intracellular side. Collectively, the results indicate that Kir channels are key modulators of pacemaker activity in newborn central pain networks. PMID:23426663

  14. 21 CFR 870.3620 - Pacemaker lead adaptor.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker lead adaptor. 870.3620 Section 870.3620...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3620 Pacemaker lead adaptor. (a) Identification. A pacemaker lead adaptor is a device used to adapt a pacemaker lead so that it...

  15. [Wide QRS tachycardia preceded by pacemaker spikes].

    Science.gov (United States)

    Romero, M; Aranda, A; Gómez, F J; Jurado, A

    2014-04-01

    The differential diagnosis and therapeutic management of wide QRS tachycardia preceded by pacemaker spike is presented. The pacemaker-mediated tachycardia, tachycardia fibrillo-flutter in patients with pacemakers, and runaway pacemakers, have a similar surface electrocardiogram, but respond to different therapeutic measures. The tachycardia response to the application of a magnet over the pacemaker could help in the differential diagnosis, and in some cases will be therapeutic, as in the case of a tachycardia-mediated pacemaker. Although these conditions are diagnosed and treated in hospitals with catheterization laboratories using the application programmer over the pacemaker, patients presenting in primary care clinic and emergency forced us to make a diagnosis and treat the haemodynamically unstable patient prior to referral. Copyright © 2012 Sociedad Española de Médicos de Atención Primaria (SEMERGEN). Publicado por Elsevier España. All rights reserved.

  16. Pacemakers and Implantable Defibrillators: MedlinePlus Health Topic

    Science.gov (United States)

    ... ClinicalTrials.gov: Pacemaker, Artificial (National Institutes of Health) Journal Articles References and abstracts from MEDLINE/PubMed (National ... Leadless Cardiac Pacemakers: The Next Evolution in Pacemaker Technology. ... on Pacemakers and Implantable Defibrillators is the National Heart, Lung, and Blood Institute Other Languages Find health information in languages other than English on Pacemakers and ...

  17. 21 CFR 870.3700 - Pacemaker programmers.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker programmers. 870.3700 Section 870.3700...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3700 Pacemaker programmers. (a) Identification. A pacemaker programmer is a device used to change noninvasively one or more of...

  18. Persistent Left Superior Vena Cava Associated with Hemiazygos Vein Draining in It and Absence of Left Brachiocephalic Vein, in a Patient with Congenital Heart Defect

    Directory of Open Access Journals (Sweden)

    Opincariu Diana

    2016-09-01

    Full Text Available Persistent left superior vena cava is an anomalous vein that derives from a malfunction of obliteration of the left common cardinal vein during intrauterine life. The diagnosis can be suggested by a dilated coronary sinus as seen in echocardiography, or other imagistic methods. Due to the lack of hemodynamic impairment, and consequently with few or no symptoms, this vascular anomaly is frequently discovered incidentally. In this brief report we present the case of a 35-year-old male known with a complex congenital cardiovascular malformation that included atrial septum defect, persistent left superior vena cava and anomalous right pulmonary vein drainage in the PLSVC, diagnosed with sinoatrial block that required pacemaker implantation. Due to the patient’s medical history, investigations to decide the best approach needed for pacemaker implantation were performed, including a thoracic CT that incidentally found additional anomalies — the hemiazygos vein draining in PLSVC and the lack of the left brachiocephalic vein.

  19. Case study thoracic radiotherapy in an elderly patient with pacemaker: The issue of pacing leads

    International Nuclear Information System (INIS)

    Kirova, Youlia M.; Menard, Jean; Chargari, Cyrus; Mazal, Alejandro; Kirov, Krassen

    2012-01-01

    To assess clinical outcome of patients with pacemaker treated with thoracic radiation therapy for T8-T9 paravertebral chloroma. A 92-year-old male patient with chloroma presenting as paravertebral painful and compressive (T8-T9) mass was referred for radiotherapy in the Department of Radiation Oncology, Institut Curie. The patient presented with cardiac dysfunction and a permanent pacemaker that had been implanted prior. The decision of Multidisciplinary Meeting was to deliver 30 Gy in 10 fractions for reducing the symptoms and controlling the tumor growth. The patient received a total dose of 30 Gy in 10 fractions using 4-field conformal radiotherapy with 20-MV photons. The dose to pacemaker was 0.1 Gy but a part of the pacing leads was in the irradiation fields. The patient was treated the first time in the presence of his radiation oncologist and an intensive care unit doctor. Moreover, the function of his pacemaker was monitored during the entire radiotherapy course. No change in pacemaker function was observed during any of the radiotherapy fractions. The radiotherapy was very well tolerated without any side effects. The function of the pacemaker was checked before and after the radiotherapy treatment by the cardiologist and no pacemaker dysfunction was observed. Although updated guidelines are needed with acceptable dose criteria for implantable cardiac devices, it is possible to treat patients with these devices and parts encroaching on the radiation field. This case report shows we were able to safely treat our patient through a multidisciplinary approach, monitoring the patient during each step of the treatment.

  20. Case study thoracic radiotherapy in an elderly patient with pacemaker: The issue of pacing leads

    Energy Technology Data Exchange (ETDEWEB)

    Kirova, Youlia M., E-mail: youlia.kirova@curie.net [Department of Radiation Oncology, Institut Curie, Paris (France); Menard, Jean; Chargari, Cyrus; Mazal, Alejandro [Department of Radiation Oncology, Institut Curie, Paris (France); Kirov, Krassen [Department of Anesthesiology and Reanimation, Institut Curie, Paris (France)

    2012-07-01

    To assess clinical outcome of patients with pacemaker treated with thoracic radiation therapy for T8-T9 paravertebral chloroma. A 92-year-old male patient with chloroma presenting as paravertebral painful and compressive (T8-T9) mass was referred for radiotherapy in the Department of Radiation Oncology, Institut Curie. The patient presented with cardiac dysfunction and a permanent pacemaker that had been implanted prior. The decision of Multidisciplinary Meeting was to deliver 30 Gy in 10 fractions for reducing the symptoms and controlling the tumor growth. The patient received a total dose of 30 Gy in 10 fractions using 4-field conformal radiotherapy with 20-MV photons. The dose to pacemaker was 0.1 Gy but a part of the pacing leads was in the irradiation fields. The patient was treated the first time in the presence of his radiation oncologist and an intensive care unit doctor. Moreover, the function of his pacemaker was monitored during the entire radiotherapy course. No change in pacemaker function was observed during any of the radiotherapy fractions. The radiotherapy was very well tolerated without any side effects. The function of the pacemaker was checked before and after the radiotherapy treatment by the cardiologist and no pacemaker dysfunction was observed. Although updated guidelines are needed with acceptable dose criteria for implantable cardiac devices, it is possible to treat patients with these devices and parts encroaching on the radiation field. This case report shows we were able to safely treat our patient through a multidisciplinary approach, monitoring the patient during each step of the treatment.

  1. On the Evolution of the Cardiac Pacemaker

    Directory of Open Access Journals (Sweden)

    Silja Burkhard

    2017-04-01

    Full Text Available The rhythmic contraction of the heart is initiated and controlled by an intrinsic pacemaker system. Cardiac contractions commence at very early embryonic stages and coordination remains crucial for survival. The underlying molecular mechanisms of pacemaker cell development and function are still not fully understood. Heart form and function show high evolutionary conservation. Even in simple contractile cardiac tubes in primitive invertebrates, cardiac function is controlled by intrinsic, autonomous pacemaker cells. Understanding the evolutionary origin and development of cardiac pacemaker cells will help us outline the important pathways and factors involved. Key patterning factors, such as the homeodomain transcription factors Nkx2.5 and Shox2, and the LIM-homeodomain transcription factor Islet-1, components of the T-box (Tbx, and bone morphogenic protein (Bmp families are well conserved. Here we compare the dominant pacemaking systems in various organisms with respect to the underlying molecular regulation. Comparative analysis of the pathways involved in patterning the pacemaker domain in an evolutionary context might help us outline a common fundamental pacemaker cell gene programme. Special focus is given to pacemaker development in zebrafish, an extensively used model for vertebrate development. Finally, we conclude with a summary of highly conserved key factors in pacemaker cell development and function.

  2. On the Evolution of the Cardiac Pacemaker

    Science.gov (United States)

    Burkhard, Silja; van Eif, Vincent; Garric, Laurence; Christoffels, Vincent M.; Bakkers, Jeroen

    2017-01-01

    The rhythmic contraction of the heart is initiated and controlled by an intrinsic pacemaker system. Cardiac contractions commence at very early embryonic stages and coordination remains crucial for survival. The underlying molecular mechanisms of pacemaker cell development and function are still not fully understood. Heart form and function show high evolutionary conservation. Even in simple contractile cardiac tubes in primitive invertebrates, cardiac function is controlled by intrinsic, autonomous pacemaker cells. Understanding the evolutionary origin and development of cardiac pacemaker cells will help us outline the important pathways and factors involved. Key patterning factors, such as the homeodomain transcription factors Nkx2.5 and Shox2, and the LIM-homeodomain transcription factor Islet-1, components of the T-box (Tbx), and bone morphogenic protein (Bmp) families are well conserved. Here we compare the dominant pacemaking systems in various organisms with respect to the underlying molecular regulation. Comparative analysis of the pathways involved in patterning the pacemaker domain in an evolutionary context might help us outline a common fundamental pacemaker cell gene programme. Special focus is given to pacemaker development in zebrafish, an extensively used model for vertebrate development. Finally, we conclude with a summary of highly conserved key factors in pacemaker cell development and function. PMID:29367536

  3. 21 CFR 870.3670 - Pacemaker charger.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker charger. 870.3670 Section 870.3670 Food... DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3670 Pacemaker charger. (a) Identification. A pacemaker charger is a device used transcutaneously to recharge the batteries of a rechargeable...

  4. What Is a Pacemaker?

    Science.gov (United States)

    ... your pacemaker. • If you work around industrial microwaves, electricity, cars or other large motors, ask your doctor about possible effects. Can I use a cell phone or microwave oven if I have a pacemaker? Microwave ovens, electric blankets, remote controls for TV and other common ...

  5. A Right-sided Aortic Arch with Kommerell's Diverticulum of the Aberrant Left Subclavian Artery Presenting with Syncope

    Directory of Open Access Journals (Sweden)

    Ming-Hsun Yang

    2009-05-01

    Full Text Available A right-sided aortic arch with an aneurysm of the aberrant subclavian artery is a rare disease. We report a case of Kommerell's diverticulum of an aberrant left subclavian artery in a patient with a right-sided aortic arch. Fewer than 50 cases have been reported in the literature. A number of operative strategies are described. Right thoracotomy provides good exposure and avoids the morbidity associated with bilateral thoracotomy or sternotomy and thoracotomy. In our patient with symptoms of dysphagia, syncope, and left subclavian steal syndrome, a left thoracotomy was used. The repair was accomplished by division of a left ligamentum arteriosum, obliteration of the Kommerell's aneurysm, and an aorto-subclavian bypass. Postoperative complications included left vocal cord palsy and Horner's syndrome. Hoarseness and left ptosis recovered spontaneously 3 months after surgery, and the patient remained symptom-free at the 1-year follow-up. We believe a left thoracotomy for direct repair of Kommerell's diverticulum is a simple and safe method without the increased morbidity found in other procedures.

  6. Pacemaker patients’ perspective and experiences in a pacemaker outpatient clinic in relation to test intervals of the pacemaker

    DEFF Research Database (Denmark)

    Lauberg, Astrid; Hansen, Tina; Pedersen, Trine Pernille Dahl

    an evident decline in quality of life regarding psychological and social aspects 6 month after the implantation in terms of cognitive function, work ability, and sexual activity. Mlynarski et al (2009) have found correlations between pacemaker implantation and anxiety and depression. Aim The aim...... the pacemaker and psychological reactions. Patients with pacemakers older than 3 months lacked communication with fellowmen. Conclusion The patients express receiving competent and professional treatment when visiting the outpatient clinic, there seems to be a discrepancy between the long test intervals...... and the critical period in which anxiety and depression may occur. Minor problems and questions may grow into fatal conditions if the patients are not offered an opportunity to discuss this with experts. Patients are not informed that it is possible to discuss problems that imply psychological topics and they do...

  7. Oral antibiotic treatment of left-sided infectious endocarditis verified by 16S-PCR

    DEFF Research Database (Denmark)

    Bruun, Louise E; Tønder, Niels; Hansen, Thomas Fritz

    2011-01-01

    Treatment of infectious endocarditis (IE) comprises intravenously administered antibiotic medications given at high doses for 4-6 weeks--sometimes even longer. Approximately 50% of patients referred to tertiary care centres require additional surgical intervention. At present there are few papers...... describing the effects of oral antibiotic treatment in IE, and only in patients with right-sided endocarditis. In this case report we present a patient with left-sided Streptococcus endocarditis successfully treated with oral antibiotic drugs....

  8. Pacemaker

    Science.gov (United States)

    ... close to a security system metal detector. Notify security staff if you have a pacemaker. Also, stay at least 2 feet away from industrial welders and electrical generators. Some medical procedures can ...

  9. Radiotherapy for breast cancer and pacemaker

    International Nuclear Information System (INIS)

    Menard, J.; Campana, F.; Bollet, M.A.; Dendale, R.; Fournier-Bidoz, N.; Marchand, V.; Mazal, A.; Fourquet, A.; Kirova, Y.M.; Kirov, K.M.; Esteve, M.

    2011-01-01

    Purpose. - Patients with permanent cardiac pacemakers occasionally require radiotherapy. Therapeutic Irradiation may cause pacemakers to malfunction due to the effects of ionizing radiation or electromagnetic interference. We studied the breast cancer patients who needed breast and/or chest wall and lymph node irradiation to assess the feasibility and tolerance in this population of patients. Patients and methods. - From November 2008 to December 2009, more than 900 patients received radiotherapy for their breast cancer in our department using megavoltage linear accelerator (X 4-6 MV and electrons). Among them, seven patients were with permanent pacemaker. All patients have been treated to the breast and chest wall and/or lymph nodes. Total dose to breast and/or chest wall was 50 Gy/25 fractions and 46 Gy/23 fractions to lymph nodes. Patients who underwent conserving surgery followed by breast irradiation were boosted when indicated to tumour bed with 16 Gy/8 fractions. All patients were monitored everyday in presence of radiation oncologist to follow the function of their pacemaker. All pacemakers were controlled before and after radiotherapy by the patients' cardiologist. Results. - Seven patients were referred in our department for postoperative breast cancer radiotherapy. Among them, only one patient was declined for radiotherapy and underwent mastectomy without radiotherapy. In four cases the pacemaker was repositioned before the beginning of radiotherapy. Six patients, aged between 48 and 84 years underwent irradiation for their breast cancer. Four patients were treated with conserving surgery followed by breast radiotherapy and two with mastectomy followed by chest wall and internal mammary chain, supra- and infra-clavicular lymph node irradiation. The dose to the pacemaker generator was kept below 2 Gy. There was no pacemaker dysfunction observed during the radiotherapy. Conclusion. - The multidisciplinary work with position change of the pacemaker before

  10. [Right-side aortic arch with aberrant left subclavian artery and Kommerell's diverticulum. A cause of vascular ring].

    Science.gov (United States)

    Tamayo-Espinosa, Tania; Erdmenger-Orellana, Julio; Becerra-Becerra, Rosario; Balderrabano-Saucedo, Norma; Segura-Standford, Begoña

    The right-side aortic arch may be associated with aberrant left subclavian artery, in some cases this artery originates from an aneurismal dilation of the aorta called Kommerell's diverticulum. A report is presented on 2 cases of vascular ring formed by a right-side aortic arch, anomalous left subclavian artery, Kommerell's diverticulum and left patent ductus arteriosus. A review the literature was also performed as regards the embryological development and the imaging methods used to help in the diagnosis of this rare vascular anomaly. Copyright © 2017 Instituto Nacional de Cardiología Ignacio Chávez. Publicado por Masson Doyma México S.A. All rights reserved.

  11. Wireless power transfer for a pacemaker application.

    Science.gov (United States)

    Vulfin, Vladimir; Sayfan-Altman, Shai; Ianconescu, Reuven

    2017-05-01

    An artificial pacemaker is a small medical device that uses electrical impulses, delivered by electrodes contracting the heart muscles, to regulate the beating of the heart. The pacemaker is implanted under the skin, and uses for many years regular non-rechargeable batteries. However, the demand for rechargeable batteries in pacemakers increased, and the aim of this work is to design an efficient charging system for pacemakers.

  12. Convergent and divergent functional connectivity patterns in patients with long-term left-sided and right-sided deafness.

    Science.gov (United States)

    Zhang, Yanyang; Mao, Zhiqi; Feng, Shiyu; Wang, Wenxin; Zhang, Jun; Yu, Xinguang

    2018-02-05

    Cortical reorganization may be induced in long-term single-sided deafness (SD); however, the influence of the deafness side on the functional changes remains poorly understood. Here, we investigated whole-brain functional connectivity patterns in long-term SD patients. The normalized voxel-based functional connectivity strength (FCS) was determined using resting-state fMRI (rs-fMRI) in 17 left-sided deafness (LD) patients, 21 right-sided deafness (RD) patients and 21 healthy controls (HCs). Relative to the HCs, both the LD and RD patients exhibited a reduction in the FCS in the ipsilateral visual cortex. However, compared to that in the HCs, a significantly higher FCS was observed in some regions in the salience and default-mode networks in the RD patients, but this FCS alternation pattern was not observed in the LD patients. A direct comparison of the two patient groups revealed a significantly increased FCS in the supplemental motor area in the LD group. Altogether, the long-term SD groups with LD and RD exhibited convergent and divergent functional connectivity patterns in whole-brain networks, providing promising evidence that the functional changes in long-term SD are highly deafness-side-dependent. Copyright © 2017 Elsevier B.V. All rights reserved.

  13. Pacemaker wires

    International Nuclear Information System (INIS)

    Fransson, S.G.

    1993-01-01

    Evaluation of pacemaker wires were performed by comparing Advanced Multiple Beam Equalization Radiography (AMBER) with conventional chest radiography. The scanning equalization technique of the AMBER unit makes it superior to conventional technique in the depiction of different structures in the mediastinum or in the pleural sinuses. So far motion artifacts have not been considered clinically important. The longer exposure time, however, may impair the assessment of pacemaker wires. The motion artifact described may not only make adequate evaluation impossible but may even give a false impression of a lead fracture. The difference between the two systems was significant. (orig.)

  14. Complications with the MICRA TPS Pacemaker System: Persistent Complete Heart Block and Late Capture Failure.

    Science.gov (United States)

    Holm, Niels; Müller, Andreas; Zbinden, Rainer

    2017-04-01

    A Medtronic MICRA transcatheter pacing system (Medtronic, Minneapolis, MN, USA) was implanted in an 86-year-old patient with sick sinus syndrome and left bundle branch block after transfemoral aortic valve implantation. During implantation she developed a persistent complete heart block due to manipulation with the large-bore delivery catheter. Two weeks later, acute pacemaker dysfunction occurred due to massive increase of pacing threshold and impedance without obvious pacemaker dislocation or myocardial perforation. Recurrent capture failure was seen with pacing output set at 5 V/1.0 ms. Hence, microdislocation or fixation of the tines in the right ventricular trabeculae has to be assumed. © 2016 Wiley Periodicals, Inc.

  15. Left-Deviating Prism Adaptation in Left Neglect Patient: Reflexions on a Negative Result

    Directory of Open Access Journals (Sweden)

    Jacques Luauté

    2012-01-01

    Full Text Available Adaptation to right-deviating prisms is a promising intervention for the rehabilitation of patients with left spatial neglect. In order to test the lateral specificity of prism adaptation on left neglect, the present study evaluated the effect of left-deviating prism on straight-ahead pointing movements and on several classical neuropsychological tests in a group of five right brain-damaged patients with left spatial neglect. A group of healthy subjects was also included for comparison purposes. After a single session of exposing simple manual pointing to left-deviating prisms, contrary to healthy controls, none of the patients showed a reliable change of the straight-ahead pointing movement in the dark. No significant modification of attentional paper-and-pencil tasks was either observed immediately or 2 hours after prism adaptation. These results suggest that the therapeutic effect of prism adaptation on left spatial neglect relies on a specific lateralized mechanism. Evidence for a directional effect for prism adaptation both in terms of the side of the visuomanual adaptation and therefore possibly in terms of the side of brain affected by the stimulation is discussed.

  16. The nuclear pacemaker: Is renewed interest warranted

    International Nuclear Information System (INIS)

    Parsonnet, V.; Berstein, A.D.; Perry, G.Y.

    1990-01-01

    From 1973 through 1987, 155 radioisotope-powered nuclear pacemakers were implanted in 132 patients at the Newark Beth Israel Medical Center. The longevity of the first 15 devices, all of which were fixed-rate (VOO) pacemakers, was significantly better than that of 15 lithium-chemistry demand (VVI) pacemakers used as control devices (p = 0.0002). Of the entire cohort of 155 nuclear pacemakers, 136 were VVI devices and 19 were VOO units. The patients with VOO pacemakers needed reoperations more often than did those with VVI pacemakers, chiefly for mode change (p less than 0.001). Power-source failure was observed in only 1 case, but 47 nuclear pacemakers were removed for other reasons, including component malfunction (15 units), mode change (12 units), high pacing thresholds (8 units) and lead or connector problems (5 units). The actuarial survival at 15 years was 99% for power sources and 82% for the entire pacing systems (pulse generators plus leads). The frequency of malignancy was similar to that of the population at large and primary tumor sites were randomly distributed. Deaths most commonly were due to cardiac causes (68%). Thus, nuclear pacemakers are safe and reliable and their greater initial cost appears to be offset by their longevity and the resulting decrease in the frequency of reoperations. It is reasonable to suggest that further use be made of long-lasting nuclear power sources for modern pacemakers and other implantable rhythm-management devices

  17. The oral cavity is not a primary source for implantable pacemaker or cardioverter defibrillator infections

    Science.gov (United States)

    2013-01-01

    Background To test the hypothesis that the oral cavity is a potential source for implantable pacemaker and cardioverter defibrillators infections, the bacterial diversity on explanted rhythm heart management devices was investigated and compared to the oral microbiome. Methods A metagenomic approach was used to analyze the bacterial diversity on the surfaces of non-infected and infected pacemakers. The DNA from surfaces swaps of 24 non-infected and 23 infected pacemaker were isolated and subjected to bacterial-specific DNA amplification, single strand conformation polymorphism- (SSCP) and sequencing analysis. Species-specific primer sets were used to analyze for any correlation between bacterial diversity on pacemakers and in the oral cavity. Results DNA of bacterial origin was detected in 21 cases on infected pacemakers and assigned to the bacterial phylotypes Staphylococcus epidermidis, Propionibacterium acnes, Staphylococcus aureus, Staphylococcus schleiferi and Stapyhlococcus. In 17 cases bacterial DNA was found on pacemakers with no clinical signs of infections. On the basis of the obtained sequence data, the phylotypes Propionibacterium acnes, Staphylococcus and an uncultured bacterium were identified. Propionibacterium acnes and Staphylococcus epidermidis were the only bacteria detected in pacemeaker (n = 25) and oral samples (n = 11). Conclusions The frequency of the coincidental detection of bacteria on infected devices and in the oral cavity is low and the detected bacteria are highly abundant colonizers of non-oral human niches. The transmission of oral bacteria to the lead or device of implantable pacemaker or cardioverter defibrillators is unlikely relevant for the pathogenesis of pacemaker or cardioverter defibrillators infections. PMID:23575037

  18. Safety of Laser Interstitial Thermal Therapy in Patients With Pacemakers.

    Science.gov (United States)

    Grewal, Sanjeet S; Gorny, Krzysztof R; Favazza, Christopher P; Watson, Robert E; Kaufmann, Timothy J; Van Gompel, Jamie J

    2018-02-10

    Laser interstitial thermal therapy (LiTT) has increasingly been used as a treatment option for medically refractory epilepsy, tumors, and radiation necrosis. The use of LiTT requires intraoperative magnetic resonance (MR) thermography. This can become an issue in patients with other implanted therapeutic devices such as pacemakers and vagal nerve stimulators due to concerns regarding increases in the specific absorption rate (SAR). This is a technical case report demonstrating a successfully and safely performed LiTT in a 1.5-T magnetic resonance imaging (MRI) in a patient with a pacemaker for mesial temporal sclerosis. An 83-yr-old gentleman who had an implanted cardiac pacemaker presented with medically intractable epilepsy and was confirmed to have mesial temporal sclerosis on imaging. Video electroencephalography demonstrated concordant ipsilateral seizures and semiology. He underwent LiTT for ablation of the mesial temporal lobe. This was performed with the below described protocol with a cardiology nurse monitoring the patient's cardiac condition and a physicist monitoring SAR, and MR imaging quality without any adverse events. This study reports on a protocol of cardiac and MR SAR to safely perform MR-guided LiTT in the setting of traditional pacemakers in patients who are not pacemaker dependent. Copyright © 2018 by the Congress of Neurological Surgeons

  19. Early experience of the compression anastomosis ring (CARTM 27) in left-sided colon resection

    Science.gov (United States)

    Lee, Jung-Yeon; Woo, Jin-Hee; Choi, Hong-Jo; Park, Ki-Jae; Roh, Young-Hoon; Kim, Ki-Han; Lee, Hak-Yoon

    2011-01-01

    AIM: To evaluate clinical validity of the compression anastomosis ring (CAR™ 27) anastomosis in left-sided colonic resection. METHODS: A non-randomized prospective data collection was performed for patients undergoing an elective left-sided colon resection, followed by an anastomosis using the CAR™ 27 between November 2009 and January 2011. Eligibility criteria of the use of the CAR™ 27 were anastomoses between the colon and at or above the intraperitoneal rectum. The primary short-term clinical endpoint, rate of anastomotic leakage, and other clinical outcomes, including intra- and postoperative complications, length of operation time and hospital stay, and the ring elimination time were evaluated. RESULTS: A total of 79 patients (male, 43; median age, 64 years) underwent an elective left-sided colon resection, followed by an anastomosis using the CAR™ 27. Colectomy was performed laparoscopically in 70 patients, in whom two patients converted to open procedure (2.9%). There was no surgical mortality. As an intraoperative complication, total disruption of the anastomosis occurred by premature enforced tension on the proximal segment of the anastomosis in one patient. The ring was removed and another new CAR™ 27 anastomosis was constructed. One patient with sigmoid colon cancer showed postoperative anastomotic leakage after 6 d postoperatively and temporary diverting ileostomy was performed. Exact date of expulsion of the ring could not be recorded because most patients were not aware that the ring had been expelled. No patients manifested clinical symptoms of anastomotic stricture. CONCLUSION: Short-term evaluation of the CAR™ 27 anastomosis in elective left colectomy suggested it to be a safe and efficacious alternative to the standard hand-sewn or stapling technique. PMID:22147979

  20. Early experience of the compression anastomosis ring (CAR™ 27) in left-sided colon resection.

    Science.gov (United States)

    Lee, Jung-Yeon; Woo, Jin-Hee; Choi, Hong-Jo; Park, Ki-Jae; Roh, Young-Hoon; Kim, Ki-Han; Lee, Hak-Yoon

    2011-11-21

    To evaluate clinical validity of the compression anastomosis ring (CAR™ 27) anastomosis in left-sided colonic resection. A non-randomized prospective data collection was performed for patients undergoing an elective left-sided colon resection, followed by an anastomosis using the CAR™ 27 between November 2009 and January 2011. Eligibility criteria of the use of the CAR™ 27 were anastomoses between the colon and at or above the intraperitoneal rectum. The primary short-term clinical endpoint, rate of anastomotic leakage, and other clinical outcomes, including intra- and postoperative complications, length of operation time and hospital stay, and the ring elimination time were evaluated. A total of 79 patients (male, 43; median age, 64 years) underwent an elective left-sided colon resection, followed by an anastomosis using the CAR™ 27. Colectomy was performed laparoscopically in 70 patients, in whom two patients converted to open procedure (2.9%). There was no surgical mortality. As an intraoperative complication, total disruption of the anastomosis occurred by premature enforced tension on the proximal segment of the anastomosis in one patient. The ring was removed and another new CAR™ 27 anastomosis was constructed. One patient with sigmoid colon cancer showed postoperative anastomotic leakage after 6 d postoperatively and temporary diverting ileostomy was performed. Exact date of expulsion of the ring could not be recorded because most patients were not aware that the ring had been expelled. No patients manifested clinical symptoms of anastomotic stricture. Short-term evaluation of the CAR™ 27 anastomosis in elective left colectomy suggested it to be a safe and efficacious alternative to the standard hand-sewn or stapling technique.

  1. The differential effects of acute right- vs. left-sided vestibular failure on brain metabolism.

    Science.gov (United States)

    Becker-Bense, Sandra; Dieterich, Marianne; Buchholz, Hans-Georg; Bartenstein, Peter; Schreckenberger, Mathias; Brandt, Thomas

    2014-07-01

    The human vestibular system is represented in the brain bilaterally, but it has functional asymmetries, i.e., a dominance of ipsilateral pathways and of the right hemisphere in right-handers. To determine if acute right- or left-sided unilateral vestibular neuritis (VN) is associated with differential patterns of brain metabolism in areas representing the vestibular network and the visual-vestibular interaction, patients with acute VN (right n = 9; left n = 13) underwent resting state (18)F-FDG PET once in the acute phase and once 3 months later after central vestibular compensation. The contrast acute vs. chronic phase showed signal differences in contralateral vestibular areas and the inverse contrast in visual cortex areas, both more pronounced in VN right. In VN left additional regions were found in the cerebellar hemispheres and vermis bilaterally, accentuated in severe cases. In general, signal changes appeared more pronounced in patients with more severe vestibular deficits. Acute phase PET data of patients compared to that of age-matched healthy controls disclosed similarities to these patterns, thus permitting the interpretation that the signal changes in vestibular temporo-parietal areas reflect signal increases, and in visual areas, signal decreases. These data imply that brain activity in the acute phase of right- and left-sided VN exhibits different compensatory patterns, i.e., the dominant ascending input is shifted from the ipsilateral to the contralateral pathways, presumably due to the missing ipsilateral vestibular input. The visual-vestibular interaction patterns were preserved, but were of different prominence in each hemisphere and more pronounced in patients with right-sided failure and more severe vestibular deficits.

  2. Guenther Tulip Filter Retrieval from a Left-sided Inferior Vena Cava

    International Nuclear Information System (INIS)

    Brountzos, Elias N.; Kaufman, John A.; Lakin, Paul L.

    2004-01-01

    Optional (retrievable) inferior cava filters (IVC) may have advantages over permanent filters in a certain subset of patients, especially in view of recent concerns about the long-term thrombotic complications of the latter. Retrieval of the Guenther Tulip Filter (GTF), an optional filter, has been reported in a total of 76 patients. We present the first description of GTF retrieval from a left-sided IVC using the right internal jugular approach

  3. 21 CFR 870.3600 - External pacemaker pulse generator.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false External pacemaker pulse generator. 870.3600... pacemaker pulse generator. (a) Identification. An external pacemaker pulse generator is a device that has a... intrinsic pacing sytem until a permanent pacemaker can be implanted, or to control irregular heartbeats in...

  4. Ionizing radiation effects on implanted pacemakers

    International Nuclear Information System (INIS)

    Holzer, J.; Aiginger, H.; Binder, W.

    1998-01-01

    Fourteen multi-programmable pacemakers and 2 intercardial defibrillators were exposed to 60 Co radiation, to 9 MeV electrons and to 6 MV and 10 MV photon radiation. The pacemakers were placed into a water phantom. The following parameters were examined: telemetry, battery, pulse frequency, pulse amplitude, and period at accumulated doses from 2 Gy to 100 Gy. It is concluded that pacemakers in CMOS/Bipolar technology and in 8μ CMOS technology should not be exposed to an absorbed dose exceeding 5 Gy, the latest generation of pacemakers in the 3μm technology will perform satisfactorily up to 70 Gy. (P.A.)

  5. Runaway pacemaker: a still existing complication and therapeutic guidelines

    DEFF Research Database (Denmark)

    Mickley, H; Andersen, C; Nielsen, L H

    1989-01-01

    Runaway pacemaker is a rare, but still existing potential lethal complication in permanent pacemakers. Within 4 1/2 years, we saw two cases of runaway pacemaker in patients with multiprogrammable, VVI pacemakers (Siemens-Elema, Model 668). In both cases a pacemaker-induced ventricular tachycardia...... pacemaker may be connected to the permanent pacing lead. Thereafter, the lead can be safely cut. As an alternative, a temporary transvenous pacing lead may be established prior to disconnecting the permanent pacing lead....

  6. Right- vs. Left-Sided Metastatic Colorectal Cancer: Differences in Tumor Biology and Bevacizumab Efficacy

    Directory of Open Access Journals (Sweden)

    Paola Ulivi

    2017-06-01

    Full Text Available There is evidence of a different response to treatment with regard to the primary tumor localization (right-sided or left-sided in patients with metastatic colorectal cancer (mCRC. We analyzed the different outcomes and biomolecular characteristics in relation to tumor localization in 122 of the 370 patients with metastatic colorectal cancer enrolled onto the phase III prospective multicenter “Italian Trial in Advanced Colorectal Cancer (ITACa”, randomized to receive first-line chemotherapy (CT or CT plus bevacizumab (CT + B. RAS and BRAF mutations; baseline expression levels of circulating vascular endothelial growth factor (VEGF, endothelial nitric oxide synthase (eNOS, cyclooxygenase-2 (COX2, ephrin type-B receptor 4 (EPHB4, hypoxia-inducible factor 1-alpha (HIF-1α, lactate dehydrogenase (LDH, and high-sensitivity C reactive protein (hs-CRP; and inflammatory indexes such as the neutrophil-to-lymphocyte ratio, platelet-lymphocyte rate and systemic immune-inflammation index were evaluated. Patients with right-sided tumors showed a longer median progression-free survival in the CT + B arm than in the CT group (12.6 vs. 9.0 months, respectively, p = 0.017. Baseline inflammatory indexes were significantly higher in left-sided tumors, whereas eNOS and EPHB4 expression was significantly higher and BRAF mutation more frequent in right-sided tumors. Our data suggest a greater efficacy of the CT + B combination in right-sided mCRC, which might be attributable to the lower inflammatory status and higher expression of pro-angiogenic factors that appear to characterize these tumors.

  7. [Successful One-lung Ventilation with a Right-sided Double-lumen Tube in a Patient with a Right Upper Tracheal Bronchus, who Underwent Left Pneumonectomy for Left Hilar Lung Cancer].

    Science.gov (United States)

    Kawagoe, Izumi; Kohchiyama, Tsukasa; Hayashida, Masakazu; Satoh, Daizoh; Suzuki, Kenji; Inada, Eiichi

    2016-06-01

    A 60-year-old male patient with left hilar lung cancer was scheduled to undergo left pneumonectomy or left sleeve lower lobectomy. Preoperative computer tomographic and bronchoscopic examinations revealed that the bronchus (B1) to the right apical segment (S1) was a tracheal bronchus (TB) originating from the trachea approximately 10 mm above the carina. Because the left main bronchus was to be dissected, a right-sided double-lumen tube (DLT) was selected to completely protect the right lung from spillage of secretions or cancer cells from the left lung. The right-sided DLT was placed so as to fit its lateral opening of the bronchial lumen to normal upper branches (B2, B3), while sacrificing ventilation of S1 with an abnormal branch (B1). However, one-lung ventilation (OLV) of the right lung could not be achieved, since a gas leakage from the opened tracheal lumen occurred, most probably due to intra-lobar micro-airway communications between S1 and S2/S3. The DLT was withdrawn until the blue bronchial cuff occluded the orifice of the TB (B1). Although the upper half of the blue bronchial cuff appeared above the tracheal carina, OLV through the two bronchial lumen openings could be achieved due to a specific, slanted doughnut shape of the blue bronchial cuff and the location of the abnormal branch (B1) approximate to the carina. Left pneumonectomy using successful OLV was completed safely without hypoxemia or hypercapnea. Our experience indicates that management of OLV for patients with a thoracheal bronchus needs special considerations of the exact location of the TB and intra-lobar micro-airway communications, in addition to types of scheduled surgical procedures.

  8. Circadian locomotor rhythms in the cricket, Gryllodes sigillatus. II. Interactions between bilaterally paired circadian pacemakers.

    Science.gov (United States)

    Ushirogawa, H; Abe, Y; Tomioka, K

    1997-10-01

    The optic lobe is essential for circadian locomotor rhythms in the cricket, Gryllodes sigillatus. We examined potential interactions between the bilaterally paired optic lobes in circadian rhythm generation. When one optic lobe was removed, the free-running period of the locomotor rhythm slightly but significantly lengthened. When exposed to light-dark cycles (LD) with 26 hr period, intact and sham operated animals were clearly entrained to the light cycle, but a large number of animals receiving unilateral optic nerve severance showed rhythm dissociation. In the dissociation, two rhythmic components appeared; one was readily entrained to the given LD and the other free-ran with a period shorter than 24 hr, and activity was expressed only when they were inphase. The period of the free-running component was significantly longer than that of the animals with a single blinded pacemaker kept in LD13:13, suggesting that the pacemaker on the intact side had some influence on the blinded pacemaker even in the dissociated state. The ratio of animals with rhythm dissociation was greater with the lower light intensity of the LD. The results suggest that the bilaterally distributed pacemakers are only weakly coupled to one another but strongly suppress the activity driven by the partner pacemaker during their subjective day. The strong suppression of activity would be advantageous to keep a stable nocturnality for this cricket living indoors.

  9. Value of Combining Left Atrial Diameter and Amino-terminal Pro-brain Natriuretic Peptide to the CHA2DS2-VASc Score for Predicting Stroke and Death in Patients with Sick Sinus Syndrome after Pacemaker Implantation.

    Science.gov (United States)

    Mo, Bin-Feng; Lu, Qiu-Fen; Lu, Shang-Biao; Xie, Yu-Quan; Feng, Xiang-Fei; Li, Yi-Gang

    2017-08-20

    The CHA2DS2-VASc score is used clinically for stroke risk stratification in patients with atrial fibrillation (AF). We sought to investigate whether the CHA2DS2-VASc score predicts stroke and death in Chinese patients with sick sinus syndrome (SSS) after pacemaker implantation and to evaluate whether the predictive power of the CHA2DS2-VASc score could be improved by combining it with left atrial diameter (LAD) and amino-terminal pro-brain natriuretic peptide (NT-proBNP). A total of 481 consecutive patients with SSS who underwent pacemaker implantation from January 2004 to December 2014 in our department were included. The CHA2DS2-VASc scores were retrospectively calculated according to the hospital medical records before pacemaker implantation. The outcome data (stroke and death) were collected by pacemaker follow-up visits and telephonic follow-up until December 31, 2015. During 2151 person-years of follow-up, 46 patients (9.6%) suffered stroke and 52 (10.8%) died. The CHA2DS2-VASc score showed a significant association with the development of stroke (hazard ratio [HR] 1.45, 95% confidence interval [CI] 1.20-1.75, Ppacemaker implantation. The addition of LAD and NT-proBNP to the CHA2DS2-VASc score improved its predictive power for stroke and death, respectively, in this patient cohort. Future prospective studies are warranted to validate the benefit of adding LAD and NT-proBNP to the CHA2DS2-VASc score for predicting stroke and death risk in non-AF populations.

  10. 21 CFR 870.3730 - Pacemaker service tools.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker service tools. 870.3730 Section 870.3730...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3730 Pacemaker service tools. (a) Identification. Pacemaker service tools are devices such as screwdrivers and Allen wrenches...

  11. Isotopic cardiac pacemaker during pregnancy. Three cases

    International Nuclear Information System (INIS)

    Laurens, P.; Gavelle, P.; Maurice, P.; Haiat, R.; Chiche, P.

    1976-01-01

    Three cases of full term pregnancies, without complications, in women with isotopic pacemakers are reported. The newborn infants were normal. In one case, pregnancy occurred in a patient who already had a pacemaker. In two cases, the pacemaker was inserted during pregnancy (at 1 and 1/2 and 5 months respectively), in the treatment of syncopal attacks due to paroxysmal atrioventricular block. This type of pacemaker using a radioactive source (plutonium 238) is, by virtue of the low degree of radiation, harmless and may be used in women of childbearing age. Under the least favourable conditions (pacemaker box in abdominal situation), the dose delivered during pregnancy (57 mrem) is approximately 20 times less than the authorized dose (1125 mrem) [fr

  12. 21 CFR 870.3690 - Pacemaker test magnet.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker test magnet. 870.3690 Section 870.3690...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3690 Pacemaker test magnet. (a) Identification. A pacemaker test magnet is a device used to test an inhibited or triggered type...

  13. Epithermal neutron beam interference with cardiac pacemakers

    International Nuclear Information System (INIS)

    Koivunoro, H.; Serén, T.; Hyvönen, H.; Kotiluoto, P.; Iivonen, P.; Auterinen, I.; Seppälä, T.; Kankaanranta, L.; Pakarinen, S.; Tenhunen, M.; Savolainen, S.

    2011-01-01

    In this paper, a phantom study was performed to evaluate the effect of an epithermal neutron beam irradiation on the cardiac pacemaker function. Severe malfunction occurred in the pacemakers after substantially lower dose from epithermal neutron irradiation than reported in the fast neutron or photon beams at the same dose rate level. In addition the pacemakers got activated, resulting in nuclides with half-lives from 25 min to 115 d. We suggest that BNCT should be administrated only after removal of the pacemaker from the vicinity of the tumor.

  14. Epithermal neutron beam interference with cardiac pacemakers

    Energy Technology Data Exchange (ETDEWEB)

    Koivunoro, H., E-mail: hanna.koivunoro@helsinki.fi [Department of Physics, P.O.B. 64, FI-00014 University of Helsinki (Finland)] [Department of Oncology, Helsinki University Central Hospital, P.O.B. 180, FIN-00029 HUS (Finland)] [Boneca Corporation, Finland, Filnland (Finland); Seren, T. [VTT Technical Research Centre of Finland (Finland); Hyvoenen, H. [Boneca Corporation, Finland, Filnland (Finland); Kotiluoto, P. [VTT Technical Research Centre of Finland (Finland); Iivonen, P. [St. Jude Medical (Finland); Auterinen, I. [VTT Technical Research Centre of Finland (Finland); Seppaelae, T.; Kankaanranta, L. [Department of Oncology, Helsinki University Central Hospital, P.O.B. 180, FIN-00029 HUS (Finland); Pakarinen, S. [Department of Cardiology, Helsinki University Central Hospital (Finland); Tenhunen, M. [Department of Oncology, Helsinki University Central Hospital, P.O.B. 180, FIN-00029 HUS (Finland); Savolainen, S. [HUS Helsinki Medical Imaging Center, Helsinki University Central Hospital (Finland)

    2011-12-15

    In this paper, a phantom study was performed to evaluate the effect of an epithermal neutron beam irradiation on the cardiac pacemaker function. Severe malfunction occurred in the pacemakers after substantially lower dose from epithermal neutron irradiation than reported in the fast neutron or photon beams at the same dose rate level. In addition the pacemakers got activated, resulting in nuclides with half-lives from 25 min to 115 d. We suggest that BNCT should be administrated only after removal of the pacemaker from the vicinity of the tumor.

  15. 21 CFR 870.3640 - Indirect pacemaker generator function analyzer.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Indirect pacemaker generator function analyzer... Indirect pacemaker generator function analyzer. (a) Identification. An indirect pacemaker generator function analyzer is an electrically powered device that is used to determine pacemaker function or...

  16. The influence of radiation therapy on cardiac pacemakers

    International Nuclear Information System (INIS)

    Coles, J.R.; Ciddor, G.S.

    1980-01-01

    The results of an investigation to determine the influence on pacemaking of ionizing radiation and electromagnetic radiation from a number of radiotherapy machines are reported. In vitro tests were carried out on unipolar cardiac pacemakers of the ventricular inhibited type. The pacemakers were largely unaffected by the environment of clinical radiotherapy machines. Ionizing radiation had no detrimental effect on the pacemakers and electromagnetic interference caused only temporary single-beat inhibition at most. With the betatron used, malfunction of the pacemakers regularly occurred whilst in their inhibited made of operation. The demand function became disabled allowing competitive asynchronous pulses to be produced

  17. Ca2+-clock-dependent pacemaking in the sinus node is impaired in mice with a cardiac specific reduction in SERCA2 abundance

    Directory of Open Access Journals (Sweden)

    Sunil Jit Ramamoorthy Jeewanlal Logantha

    2016-06-01

    Full Text Available Background: The sarcoplasmic reticulum Ca2+-ATPase (SERCA2 pump is an important component of the Ca2+-clock pacemaker mechanism that provides robustness and flexibility to sinus node pacemaking. We have developed transgenic mice with reduced cardiac SERCA2 abundance (Serca2 KO as a model for investigating SERCA2’s role in sinus node pacemaking.Methods and Results: In Serca2 KO mice, ventricular SERCA2a protein content measured by Western blotting was 75% (P70% Serca2 downregulation.Conclusions: Serca2 KO mice show a disrupted Ca2+-clock-dependent pacemaker mechanism contributing to impaired sinus node and atrioventricular node function.

  18. Comparison of biatrial and left atrial radiofrequency ablation in patients with heart valves diseases

    Directory of Open Access Journals (Sweden)

    А. В. Богачев-прокофьев

    2015-10-01

    Full Text Available We analysed results of left atrial and biatrial anatomic scheme of radiofrequency (RF ablation of the atria in patients with valvular heart. From 2007 to 2011, RF ablation procedure performed 283 patients with atrial fibrillation and heart valve disease. Patients were divided into two groups - group I performed a complete circuit biatrial procedure (procedure maze IV - 131 patients, in group II only ablation of the left atrium -152 patients. Most of the patients had long-term persistent AF. The main problem of early postoperative period is dysfunction of pacemaker complex, which observed in 64.9% and 50.7% patients for group I and II, respectively (p = 0.031. Permanent dysfunction of pacemaker complex required implantation of permanent pacemaker (9.2% and 4.6% of cases for group I and II, respectively, p = 0,023. Predictors of dysfunction of the pacemaker complex were biatrial execution method and duration of AF. Freedom from AF and AT during 36 months after surgery was 80.4% for patients in group I and 78.5% for patients in group II, with no statistically significant difference (log-rank test, p = 0.621. Atrial transport function has not revealed that the active fraction of the right atrium was 29.8% higher than in group II in the early postoperative phase and by 17.9% in the long term, with no difference between groups for the left atrium has been received.

  19. Echocardiographic 3D-guided 2D planimetry in quantifying left-sided valvular heart disease.

    Science.gov (United States)

    Argulian, Edgar; Seetharam, Karthik

    2018-02-08

    Echocardiographic 3D-guided 2D planimetry can improve the accuracy of valvular disease assessment. Acquisition of 3D pyramidal dataset allows subsequent multiplanar reconstruction with accurate orthogonal plane alignment to obtain the correct borders of an anatomic orifice or flow area. Studies examining the 3D-guided 2D planimetry approach in left-sided valvular heart disease were identified and reviewed. The strongest evidence exists for estimating mitral valve area in patients with rheumatic mitral valve stenosis and vena contracta area in patients with mitral regurgitation (both primary and secondary). 3D-guided approach showed excellent feasibility and reproducibility in most studies, as well as time efficiency and good correlation with reference and comparator methods. Therefore, 3D-guided 2D planimetry can be used as an important clinical tool in quantifying left-sided valvular heart disease, especially mitral valve disorders. © 2018 Wiley Periodicals, Inc.

  20. [Left-sided native valve endocarditis by coagulase-negative staphylococci: an emerging disease].

    Science.gov (United States)

    Haro, Juan Luis; Lomas, José M; Plata, Antonio; Ruiz, Josefa; Gálvez, Juan; de la Torre, Javier; Hidalgo-Tenorio, Carmen; Reguera, José M; Márquez, Manuel; Martínez-Marcos, Francisco; de Alarcón, Arístides

    2008-05-01

    To describe the epidemiological, clinical, and prognostic characteristics of patients with left-sided native valve endocarditis (LNVE) caused by coagulase-negative staphylococci (CoNS). Prospective multicenter study of endocarditis cases reported in the Andalusian Cohort for the Study of Cardiovascular Infections between 1984 and 2005. Among 470 cases of LNVE, 39 (8.3%) were caused by CoNS, a number indicating a 30% increase in the incidence of this infection over the last decade. The mean age of affected patients was 58.32 +/- 15 years and 27 (69.2%) were men. Twenty-one patients (53.8%) had previous known valve disease and half the episodes were considered nosocomial (90% of them from vascular procedures). Median time interval from the onset of symptoms to diagnosis was 14 days (range: 1-120). Renal failure (21 cases, 53.8%), intracardiac damage (11 cases, 28.2%), and central nervous system involvement (10 cases, 25.6%) were the most frequent complications. There were only 3 cases (7.7%) of septic shock. Surgery was performed in 18 patients (46.2%). Nine patients (23.1%) died, overall. Factors associated with higher mortality in the univariate analysis were acute renal failure (P = 0.023), left-sided ventricular failure (P = 0.047), and time prior to diagnosis less than 21 days (P = 0.018). As compared to LNVE due to other microorganisms, the patients were older (P = 0.018), had experienced previous nosocomial manipulation as the source of bacteremia (P < 0.001), and developed acute renal failure more frequently (P = 0.001). Mortality of LNVE due to CoNS was lower than mortality in Staphylococcus aureus infection, but higher than in Streptococcus viridans infection. Left-sided native valve endocarditis due to CoNS is now increasing because of the ageing of the population. This implies more frequent invasive procedures (mainly vascular) as a consequence of the concomitant disease. Nonetheless, the mortality associated with LNVE due to CoNS does not seem to be

  1. [Comparison of ablation of left-sided accessory pathway by atrial septal and retrograde arterial approach].

    Science.gov (United States)

    Zhu, J G; Bao, Z Y; Gu, X

    2017-03-07

    Objective: To compare the advantages and disadvantages of radiofrequency ablation of left-sided accessory pathways by via atrial septal approach with retrograde through aortic approach. Methods: A total of 184 patients of left-side accessory pathways were treated in Taizhou People's Hospital and the Subei People's Hospital from March 2012 to August 2015.A total of 103 cases were treated by aortic retrograde approach as through arterial group, 81 cases were treated by punctured atrial septal to left atrial for mapping and ablation as through atrial septal group.Comparison of ablation procedure time, total and pathways of different parts(subgroup) at instant success and relapse rates, safety (serious complications), and statistics other complications in operation and postoperative. Results: Through arterial group and through atrial septal group were no significant difference ( P >0.05) in the ablation procedure time((25±18 ) vs (22±15)min ), instant success(98.1% vs 97.5%) and relapse rates(1.0% vs 1.2%), security(1 vs 0 case). There was no statistical difference in septal part subgroups (all P >0.05) in the ablation procedure time((22±18)vs (25±19)min), instant success(91.7% vs 89.9 %) and relapse rates(0 vs 11.1%); posterior wall subgroup had no statistical difference in the ablation procedure time((18±15)vs (16±12)min), instant success(100% vs 100 %) and relapse rates(0 vs 0)(all P >0.05); side wall subgroup had no statistical difference in the ablation procedure time((29±20)vs (21±18) min), instant success (98.3% vs 98.1%)and relapse rates(1.7% vs 0%)(all P >0.05). Conclusion: Ablation of left-sided accessory pathways by transseptal approach and transaortic approach has no statistical difference in the procedure time, instant success and relapse rates, security.In a particular case, there is a certain complementarity between the two methods.

  2. Proton Beam Therapy Interference With Implanted Cardiac Pacemakers

    International Nuclear Information System (INIS)

    Oshiro, Yoshiko; Sugahara, Shinji; Noma, Mio; Sato, Masato; Sakakibara, Yuzuru; Sakae, Takeji; Hayashi, Yasutaka; Nakayama, Hidetsugu; Tsuboi, Koji; Fukumitsu, Nobuyoshi; Kanemoto, Ayae; Hashimoto, Takayuki; Tokuuye, Koichi

    2008-01-01

    Purpose: To investigate the effect of proton beam therapy (PBT) on implanted cardiac pacemaker function. Methods and Materials: After a phantom study confirmed the safety of PBT in patients with cardiac pacemakers, we treated 8 patients with implanted pacemakers using PBT to a total tumor dose of 33-77 gray equivalents (GyE) in dose fractions of 2.2-6.6 GyE. The combined total number of PBT sessions was 127. Although all pulse generators remained outside the treatment field, 4 patients had pacing leads in the radiation field. All patients were monitored by means of electrocardiogram during treatment, and pacemakers were routinely examined before and after PBT. Results: The phantom study showed no effect of neutron scatter on pacemaker generators. In the study, changes in heart rate occurred three times (2.4%) in 2 patients. However, these patients remained completely asymptomatic throughout the PBT course. Conclusions: PBT can result in pacemaker malfunctions that manifest as changes in pulse rate and pulse patterns. Therefore, patients with cardiac pacemakers should be monitored by means of electrocardiogram during PBT

  3. [Ventricular tachycardia in a patient with rate-responsive cardiac pacemaker].

    Science.gov (United States)

    Himbert, C; Lascault, G; Tonet, J; Coutte, R; Busquet, P; Frank, R; Grosgogeat, Y

    1992-11-01

    The authors report a case of syncopal ventricular tachycardia in a patient with a respiratory-dependent rate responsive pacemaker, followed-up for valvular heart disease with severe left ventricular dysfunction and sustained atrial and ventricular arrhythmias. The introduction of low dose betablocker therapy with reinforcement of the treatment of cardiac failure controlled the ventricular arrhythmia, after suppression of the data responsive function had been shown to be ineffective. The authors discuss the role of the rate responsive function in the triggering of the ventricular tachycardias.

  4. Chylous ascites as a complication of left sided robot-assisted laparoscopic partial nephrectomy

    Directory of Open Access Journals (Sweden)

    Gaurav Pahouja

    2016-10-01

    Full Text Available Objective: The aim of the study was to present a case series of the sparsely reported complication of chylous ascites (CA after left sided robot-assisted laparoscopic partial nephrectomy (RALPN, identify possible risk factors for the development of postoperative CA, and explore current recommendations for identification, management and prevention of CA. Material and methods: A retrospective review of patients that were treated with a RALPN during a one year time period (August 2012 to August 2013 by one surgeon at our institution was conducted. A total of 12 patients were included in the study. Demographics, tumor characteristics, and perioperative outcomes were assessed. Results: Three patients in the study experienced postoperative CA. All three patients had left sided surgery. The initial clinical suspicion for CA was raised due to complaints of abdominal pain with increased milky appearance of JP fluid. JP triglycerides were elevated in all three patients. The patients responded to conservative measures, with two patients treated with medium chain triglyceride diets and one patient treated with total parenteral nutrition (TPN. Among the patients treated with RALPN, the group that was diagnosed with postoperative CA (CA group was found to have a statistically significant lower average body mass index (BMI as compared to the group that did not have CA (non-CA group (24.67 kg/m2 in the CA group versus 31.77 kg/m2 in the non-CA group; P = 0.026. Other demographic data, tumor characteristics, and perioperative outcomes were similar in both groups. Conclusions: CA as a result of RALPN is a newly reported and rare postoperative complication. As utilization of RALPN continues to increase, urologists should be aware of this possible complication and be adept at diagnosing and managing CA. We suggest that left sided retroperitoneal surgery and a lower BMI preoperatively be considered risk factors for developing this complication.

  5. Pacemakers and implantable cardioverter defibrillators - general and anesthetic considerations

    Directory of Open Access Journals (Sweden)

    Amy G. Rapsang

    2014-06-01

    Full Text Available A pacemaking system consists of an impulse generator and lead or leads to carry the electrical impulse to the patient's heart. Pacemaker and implantable cardioverter defibrillator codes were made to describe the type of pacemaker or implantable cardioverter defibrillator implanted. Indications for pacing and implantable cardioverter defibrillator implantation were given by the American College of Cardiologists. Certain pacemakers have magnet-operated reed switches incorporated; however, magnet application can have serious adverse effects; hence, devices should be considered programmable unless known otherwise. When a device patient undergoes any procedure (with or without anesthesia, special precautions have to be observed including a focused history/physical examination, interrogation of pacemaker before and after the procedure, emergency drugs/temporary pacing and defibrillation, reprogramming of pacemaker and disabling certain pacemaker functions if required, monitoring of electrolyte and metabolic disturbance and avoiding certain drugs and equipments that can interfere with pacemaker function. If unanticipated device interactions are found, consider discontinuation of the procedure until the source of interference can be eliminated or managed and all corrective measures should be taken to ensure proper pacemaker function should be done. Post procedure, the cardiac rate and rhythm should be monitored continuously and emergency drugs and equipments should be kept ready and consultation with a cardiologist or a pacemaker-implantable cardioverter defibrillator service may be necessary.

  6. Analysis of the asymmetrically expressed Ablim1 locus reveals existence of a lateral plate Nodal-independent left sided signal and an early, left-right independent role for nodal flow

    Directory of Open Access Journals (Sweden)

    Hilton Helen

    2010-05-01

    Full Text Available Abstract Background Vertebrates show clear asymmetry in left-right (L-R patterning of their organs and associated vasculature. During mammalian development a cilia driven leftwards flow of liquid leads to the left-sided expression of Nodal, which in turn activates asymmetric expression of the transcription factor Pitx2. While Pitx2 asymmetry drives many aspects of asymmetric morphogenesis, it is clear from published data that additional asymmetrically expressed loci must exist. Results A L-R expression screen identified the cytoskeletally-associated gene, actin binding lim protein 1 (Ablim1, as asymmetrically expressed in both the node and left lateral plate mesoderm (LPM. LPM expression closely mirrors that of Nodal. Significantly, Ablim1 LPM asymmetry was detected in the absence of detectable Nodal. In the node, Ablim1 was initially expressed symmetrically across the entire structure, resolving to give a peri-nodal ring at the headfold stage in a flow and Pkd2-dependent manner. The peri-nodal ring of Ablim1 expression became asymmetric by the mid-headfold stage, showing stronger right than left-sided expression. Node asymmetry became more apparent as development proceeded; expression retreated in an anticlockwise direction, disappearing first from the left anterior node. Indeed, at early somite stages Ablim1 shows a unique asymmetric expression pattern, in the left lateral plate and to the right side of the node. Conclusion Left LPM Ablim1 is expressed in the absence of detectable LPM Nodal, clearly revealing existence of a Pitx2 and Nodal-independent left-sided signal in mammals. At the node, a previously unrecognised action of early nodal flow and Pkd2 activity, within the pit of the node, influences gene expression in a symmetric manner. Subsequent Ablim1 expression in the peri-nodal ring reveals a very early indication of L-R asymmetry. Ablim1 expression analysis at the node acts as an indicator of nodal flow. Together these results make

  7. The influence of right ventricular apical pacing on left atrial volume in patients with normal left ventricular function

    Directory of Open Access Journals (Sweden)

    AR Moaref1

    2008-03-01

    Full Text Available Background: Right ventricular apical (RVA pacing has been reported to induce several deleterious effects particularly in the presence of structural heart disease but can also involve patients with normal left ventricular (LV function. Left atrial (LA enlargement is one of these effects, but the majority of studies have measured LA dimension rather than volume.Objective: The present prospective study was designed to assess the effect of RVA pacing on LA volume in patients with normal LV function.Patients and Methods: The study comprised 41 consecutive patients with LV ejection fraction ≥ 45% and LV end diastolic dimension ≤ 56 mm who underwent single-or dual- chamber pacemaker implantation in RVA and followed for LA volume measurement and pacemaker analysis at least during the ensuing 4.2 months. Results: In all, 21 patients were excluded from the study due to five spontaneous wide QRS complex (≥120msec, one recent acute coronary syndrome,one significant valvular heart disease, three pacing frequency <90%, eight death or losing follow up in three cases. In remaining 20 patients, LA volume ragned from 21 to 54 mm3 with mean of 37.3±9.7 mm3 prior to pacemaker implantation that increased to 31 to 103 mm3 (54.3±17.0 during follow-up (P<0.001.Conclusion: RVA pacing might lead to an increase in LA volume even in patients with normal LV function.

  8. 1978 Pacemaker Newspaper Awards: What Makes a Pacemaker?

    Science.gov (United States)

    Brasler, Wayne

    1979-01-01

    Lists the nine high school and college newspapers, and the one newsmagazine, that won Pacemaker Awards in 1978; discusses characteristics that make each of them outstanding, and provides reproductions of a front page from each publication. (GT)

  9. Biofilm on artificial pacemaker: fiction or reality?

    Science.gov (United States)

    Santos, Ana Paula Azevedo; Watanabe, Evandro; Andrade, Denise de

    2011-11-01

    Cardiac pacing through cardiac pacemaker is one of the most promising alternatives in the treatment of arrhythmias, but it can cause reactions natural or complex reactions, either early or late. This study aimed to describe the scientific evidence on the risk of infection and biofilm formation associated with cardiac pacemaker. This is a study of integrative literature review. It included 14 publications classified into three thematic categories: diagnosis (microbiological and/or clinical), complications and therapy of infections. Staphylococcus epidermidis and Staphylococcus aureus were the microorganisms most frequently isolated. It was not possible to determine the incidence of infection associated with pacemakers, since the studies were generally of prevalence. In terms of therapy, the complete removal of pacemakers stood out, especially in cases of suspected biofilm. Still controversial is the use of systemic antibiotic prophylaxis in reducing the incidence of infection associated with implantation of a pacemaker.

  10. 21 CFR 870.3720 - Pacemaker electrode function tester.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker electrode function tester. 870.3720... (CONTINUED) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3720 Pacemaker electrode function tester. (a) Identification. A pacemaker electrode function tester is a device which is...

  11. 21 CFR 870.3630 - Pacemaker generator function analyzer.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker generator function analyzer. 870.3630... (CONTINUED) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3630 Pacemaker generator function analyzer. (a) Identification. A pacemaker generator function analyzer is a device that is...

  12. Synchronization of Coupled Neurons Controlled by a Pacemaker

    International Nuclear Information System (INIS)

    Li Mei-Sheng; Zhang Hong-Hui; Zhao Yong; Shi Xia

    2011-01-01

    We investigate synchronization of Hindmarsh—Rose neurons with gap junctions under the control of a pacemaker. In a ring Hindmarsh—Rose neuronal network, the coupled neurons with the pacemaker can occur in synchronization more easily than those without the pacemaker. Furthermore, the pacemaker can induce phase synchronization or nearly-complete synchronization of nonidentical neurons. This synchronization can occur more easily when time delay is considered. Theses results can be helpful to understand the activities of the real neuronal system. (general)

  13. Gender specific pattern of left ventricular cardiac adaptation to ...

    African Journals Online (AJOL)

    Background: Cardiac adaptation to hypertension and obesity may be related to many factors such as race, gender and haemodynamic status. Some gender specific associations with left ventricular structure and function have been described among Caucasians. Objectives: To describe the sex specific pattern of left ...

  14. The Optimal Approach for Laparoscopic Adrenalectomy through Mono Port regarding Left or Right Sides: A Comparative Study

    Directory of Open Access Journals (Sweden)

    Wooseok Byon

    2014-01-01

    Full Text Available Introduction. Several studies have shown the feasibility and safety of both transperitoneal and posterior retroperitoneal approaches for single incision laparoscopic adrenalectomy, but none have compared the outcomes according to the left- or right-sided location of the adrenal glands. Materials and Methods. From 2009 to 2013, 89 patients who received LAMP (laparoscopic adrenalectomy through mono port were analyzed. The surgical outcomes attained using the transperitoneal approach (TPA and posterior retroperitoneal approach (PRA were analyzed and compared. Results and Discussion. On the right side, no significant differences were found between the LAMP-TPA and LAMP-PRA groups in terms of patient characteristics and clinicopathological data. However, outcomes differed in which LAMP-PRA group had a statistically significant shorter mean operative time (84.13 ± 41.47 min versus 116.84 ± 33.17 min; P=0.038, time of first oral intake (1.00 ± 0.00 days versus 1.21 ± 0.42 days; P=0.042, and length of hospitalization (2.17 ± 0.389 days versus 3.68 ± 1.38 days; P≤0.001, whereas in left-sided adrenalectomies LAMP-TPA had a statistically significant shorter mean operative time (83.85 ± 27.72 min versus 110.95 ± 29.31 min; P=0.002. Conclusions. We report that LAMP-PRA is more appropriate for right-sided laparoscopic adrenalectomies due to anatomical characteristics and better surgical outcomes. For left-sided laparoscopic adrenalectomies, however, we propose LAMP-TPA as a more suitable method.

  15. Spiral–pacemaker interactions in a mathematical model of excitable medium

    International Nuclear Information System (INIS)

    Shajahan, T K; Borek, Bartłomiej; Shrier, Alvin; Glass, Leon

    2013-01-01

    Interactions of a spiral wave with a pacemaker is studied in a mathematical model of two dimensional excitable medium. Faster pacemakers emitting target waves can abolish spirals by driving them to the border of the medium. Our study shows that a slower pacemaker can modify spiral wave behavior by changing the motion of the spiral core. We analyze the dynamics of the spiral wave near the spiral core and away from the core as a function of size and period of the pacemaker. The pacemaker can cause the spiral wave to drift towards it, and either speed up or slow down the reentrant activity. Furthermore, the drift induced by the pacemaker can result in irregular or quasiperiodic dynamics even at sites away from the pacemaker. These results highlight the influence of pacemakers on complex spiral wave dynamics. (paper)

  16. 21 CFR 870.3610 - Implantable pacemaker pulse generator.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Implantable pacemaker pulse generator. 870.3610 Section 870.3610 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES... pacemaker pulse generator. (a) Identification. An implantable pacemaker pulse generator is a device that has...

  17. 21 CFR 870.3650 - Pacemaker polymeric mesh bag.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker polymeric mesh bag. 870.3650 Section 870...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3650 Pacemaker polymeric mesh bag. (a) Identification. A pacemaker polymeric mesh bag is an implanted device used to hold a...

  18. Radiation therapy in patients with electric cardiac pacemakers

    International Nuclear Information System (INIS)

    Bisping, H.J.; Stockberg, H.; Meyer, J.; Frik, W.; Technische Hochschule Aachen

    1977-01-01

    In the course of radiation therapy and connected diagnostic measures ionizing radiation and other sources of disturbance may interfere with the function of permanent pacemakers. The conditions of such hazards are investigated in theory and practice making allowance for the different susceptibility to trouble of various models of permanent pacemakers. It appears that no extension of long-term follow-up of the cardiac pacemaker's function is needed with regard to possible late effects of ionizing radiation, but that the follow-up of pacemaker-patients during their first period of treatment should not be neglected, since other sources of electronic interference may be present. Routine checks at radiotherapy installations should also include possible sources of disturbance to electronic pacemakers. (orig.) [de

  19. 21 CFR 870.3680 - Cardiovascular permanent or temporary pacemaker electrode.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Cardiovascular permanent or temporary pacemaker... § 870.3680 Cardiovascular permanent or temporary pacemaker electrode. (a) Temporary pacemaker electrode—(1) Identification. A temporary pacemaker electrode is a device consisting of flexible insulated...

  20. VMAT techniques for lymph node-positive left sided breast cancer

    International Nuclear Information System (INIS)

    Pasler, Marlies; Lutterbach, Johannes; Bjoernsgard, Mari; Reichmann, Ursula; Bartelt, Susanne; Georg, Dietmar; Medical Univ. Vienna

    2015-01-01

    To investigate the plan quality of two different volumetric modulated arc therapy (VMAT) techniques for lymph node-positive left-sided breast cancer. Two VMAT plans were generated for 10 lymph node-positive left-sided breast cancer patients: one plan using one single segment of a full rotation, typically an arc segment of 230 (1s-VMAT); and a second plan consisting of 2 small tangential arc segments of about 50 (2s-VMAT). For plan comparison, various dose and dose volume metrics (D mean , D 98% , D 2% for target volumes, D 2% , D mean and V x% for organs at risk (OAR)) were evaluated. Both techniques fulfilled both clinical target dose and OAR goals. 1s-VMAT achieved a slightly better homogeneity and better target coverage (D 2% = 54.2 ± 0.7 Gy, D 98% = 30.3 ± 1.8 Gy) compared to 2s-VMAT (D 2% = 55.0 ± 1.1 Gy, D 98% = 29.9 ± 1.7 Gy). For geometrical reasons, OAR sparing was noticeable but not significant better using 2s-VMAT, particularly heart and contralateral breast. The heart received a mean dose of 4.4 ± 0.8 Gy using 1s-VMAT and 3.3 ± 1.0 Gy using 2s-VMAT; the contralateral breast received 1.5 ± 0.3 Gy and 0.9 ± 0.3 Gy, respectively. A VMAT technique based on two small tangential arc segments enables improved OAR sparing; the differences between the two techniques in target coverage and homogeneity are minor. Patient age and -anatomy must be considered for each individual case when deciding which technique to be used.

  1. Impact of Pacemaker Lead Characteristics on Pacemaker Related Infection and Heart Perforation: A Nationwide Population-Based Cohort Study.

    Science.gov (United States)

    Lin, Yu-Sheng; Chen, Tien-Hsing; Hung, Sheng-Ping; Chen, Dong Yi; Mao, Chun-Tai; Tsai, Ming-Lung; Chang, Shih-Tai; Wang, Chun-Chieh; Wen, Ming-Shien; Chen, Mien-Cheng

    2015-01-01

    Several risk factors for pacemaker (PM) related complications have been reported. However, no study has investigated the impact of lead characteristics on pacemaker-related complications. Patients who received a new pacemaker implant from January 1997 to December 2011 were selected from the Taiwan National Health Insurance Database. This population was grouped according to the pacemaker lead characteristics in terms of fixation and insulation. The impact of the characteristics of leads on early heart perforation was analyzed by multivariable logistic regression analysis, while the impact of the lead characteristics on early and late infection and late heart perforation over a three-year period were analyzed using Cox regression. This study included 36,104 patients with a mean age of 73.4±12.5 years. In terms of both early and late heart perforations, there were no significant differences between groups across the different types of fixation and insulations. In the multivariable Cox regression analysis, the pacemaker-related infection rate was significantly lower in the active fixation only group compared to either the both fixation (OR, 0.23; 95% CI, 0.07-0.80; P = 0.020) or the passive fixation group (OR, 0.26; 95% CI, 0.08-0.83; P = 0.023). There was no difference in heart perforation between active and passive fixation leads. Active fixation leads were associated with reduced risk of pacemaker-related infection.

  2. Impact of Pacemaker Lead Characteristics on Pacemaker Related Infection and Heart Perforation: A Nationwide Population-Based Cohort Study.

    Directory of Open Access Journals (Sweden)

    Yu-Sheng Lin

    Full Text Available Several risk factors for pacemaker (PM related complications have been reported. However, no study has investigated the impact of lead characteristics on pacemaker-related complications.Patients who received a new pacemaker implant from January 1997 to December 2011 were selected from the Taiwan National Health Insurance Database. This population was grouped according to the pacemaker lead characteristics in terms of fixation and insulation. The impact of the characteristics of leads on early heart perforation was analyzed by multivariable logistic regression analysis, while the impact of the lead characteristics on early and late infection and late heart perforation over a three-year period were analyzed using Cox regression. This study included 36,104 patients with a mean age of 73.4±12.5 years. In terms of both early and late heart perforations, there were no significant differences between groups across the different types of fixation and insulations. In the multivariable Cox regression analysis, the pacemaker-related infection rate was significantly lower in the active fixation only group compared to either the both fixation (OR, 0.23; 95% CI, 0.07-0.80; P = 0.020 or the passive fixation group (OR, 0.26; 95% CI, 0.08-0.83; P = 0.023.There was no difference in heart perforation between active and passive fixation leads. Active fixation leads were associated with reduced risk of pacemaker-related infection.

  3. Left-Sided Catamenial Pneumothorax with Thoracic Endometriosis and Bullae in the Alveolar Wall.

    Science.gov (United States)

    Takahashi, Ryo; Kurihara, Masatoshi; Mizobuchi, Teruaki; Ebana, Hiroki; Yamanaka, Sumitaka

    2017-04-20

    Catamenial pneumothorax (CP) is generally caused by intraperitoneal air leaking from the uterus into the thoracic cavity via a defect in the endometrial tissue of the diaphragm and is usually detected in the right thorax. We report a case of left-sided CP caused by endometriosis in the visceral pleura and with no abnormal findings in the diaphragm. A 33-year-old female patient presented at the end of a course of low-dose contraceptive pills for pelvic endometriosis, with spontaneous pneumothorax in the left chest. Chest CT revealed a bulla in the left upper lung lobe. The patient underwent partial resection of the lung. Immunohistochemistry confirmed the presence of endometrial stromal tissue in the visceral pleura and confirmed this as the cause of pneumothorax since there were no observable abnormalities in the diaphragm. This case suggests that immunohistochemical examination of patients with spontaneous pneumothorax can detect alternative endometrial lesions.

  4. Two Young Women with Left-sided Pneumothorax Due to Thoracic Endometriosis.

    Science.gov (United States)

    Yukumi, Shungo; Suzuki, Hideaki; Morimoto, Masamitsu; Shigematsu, Hisayuki; Okazaki, Mikio; Abe, Masahiro; Kitazawa, Sohei; Nakamura, Kenji; Sano, Yoshifumi

    Pneumothorax associated with thoracic endometriosis (TE) generally occurs in women around 30 years old and it usually affects the right pleural cavity. We herein report two cases of TE associated with left-sided pneumothorax in young women. The prevalence of TE in younger patients may be underestimated if these cases are treated as spontaneous pneumothorax. Pneumothorax occurring in younger patients has not been reported to show laterality. TE-related or catamenial pneumothorax in young women must therefore represent a different clinical entity from the condition seen in older patients.

  5. Comparison of oncological outcomes of right-sided colon cancer versus left-sided colon cancer after curative resection: Which side is better outcome?

    Science.gov (United States)

    Lim, Dae Ro; Kuk, Jung Kul; Kim, Taehyung; Shin, Eung Jin

    2017-10-01

    There are embryological origins, anatomical, histological, genetic, and immunological differences between right-sided colon cancer (RCC) and left-sided colon cancer (LCC). Many studies have sought to determine the survival and prognosis according to tumor location. This study aimed to analyze outcomes between RCC and LCC. Between January 2000 and December 2012, data on 414 patients who underwent curative resection for RCC and LCC were retrieved from a retrospective database. Propensity score matching (1:1) was performed and RCC was identified in 207 and LCC in 207 patients. On average, RCC exhibited a more advanced N stage, increased tumor size, more frequently poorly differentiated tumors, more harvested lymph nodes, and more positivity of lymphovascular invasion than LCC. With a median follow-up of 66.7 months, the 5-year overall survival (OS) rates for RCC and LCC were 82.1% and 88.7%, respectively, (P cancers, the DFS rates were 61.1% (RCC) and 81.9% (LCC; P colon cancer is needed.

  6. Wenckebach upper rate response in single chamber pacemaker.

    Science.gov (United States)

    Barold, S S

    2000-07-01

    The Medtronic Minix pacemaker during normal function in the VVT mode was found to exhibit a Wenckenbach upper rate response similar to that of dual chamber devices. This behavior occurred only when the upper rate interval was longer than the pacemaker refractory period. In a single chamber device this response may simulate pacemaker malfunction.

  7. Intermittent pacemaker dysfunction caused by digital mobile telephones.

    Science.gov (United States)

    Naegeli, B; Osswald, S; Deola, M; Burkart, F

    1996-05-01

    This study was designed to evaluate possible interactions between digital mobile telephones and implanted pacemakers. Electromagnetic fields may interfere with normal pacemaker function. Development of bipolar sensing leads and modern noise filtering techniques have lessened this problem. However, it remains unclear whether these features also protect from high frequency noise arising from digital cellular phones. In 39 patients with an implanted pacemaker (14 dual-chamber [DDD], 8 atrial-synchronized ventricular-inhibited [VDD(R)] and 17 ventricular-inhibited [VVI(R)] pacemakers), four mobile phones with different levels of power output (2 and 8 W) were tested in the standby, dialing and operating mode. During continuous electrocardiographic monitoring, 672 tests were performed in each mode with the phones positioned over the pulse generator, the atrial and the ventricular electrode tip. The tests were carried out at different sensitivity settings and, where possible, in the unipolar and bipolar pacing modes as well. In 7 (18%) of 39 patients, a reproducible interference was induced during 26 (3.9%) of 672 tests with the operating phones in close proximity (phone and at maximal sensitivity of the pacemakers (maximal vs. nominal sensitivity, 6% vs. 1.8% positive test results, p = 0.009). When the bipolar and unipolar pacing modes were compared in the same patients, ventricular inhibition was induced only in the unipolar mode (12.5% positive test results, p = 0.0003). Digital mobile phones in close proximity to implanted pacemakers may cause intermittent pacemaker dysfunction with inappropriate ventricular tracking and potentially dangerous pacemaker inhibition.

  8. Waiting for a pacemaker

    DEFF Research Database (Denmark)

    Risgaard, Bjarke; Elming, Hanne; Jensen, Gunnar

    2012-01-01

    AIMS: To determine waiting period-related morbidity, mortality, and adverse events in acute patients waiting for a permanent pacemaker (PPM).METHODS AND RESULTS: A retrospective chart review of all PPM implantations in Region Zealand, Denmark, in 2009 was conducted. Patients were excluded...... at least one adverse event during the waiting period. The present study indicates that a waiting period is dangerous as it is associated with an increased risk of adverse events. Acute PPMs should be implanted with a 24-h pacemaker implantation service capacity....

  9. Left-sided appendicitis: review of 95 published cases and a case report.

    Science.gov (United States)

    Akbulut, Sami; Ulku, Abdullah; Senol, Ayhan; Tas, Mahmut; Yagmur, Yusuf

    2010-11-28

    To give an overview of the literature on left-sided acute appendicitis (LSAA) associated with situs inversus totalis (SIT) and midgut malrotation (MM). We present a new case of LSAA with SIT and a literature review of studies published in the English language on LSAA, accessed via PubMed and Google Scholar databases. Ninety-five published cases of LSAA were evaluated and a 25-year-old female, who presented to our clinic with left lower abdominal pain caused by LSAA, is reported. In the reviewed literature, fifty-seven patients were male and 38 were female with an age range of 8 to 82 years and a median age of 29.1 ± 15.9 years. Sixty-six patients had SIT, 23 had MM, three had cecal malrotation, and two had a previously unnoted congenital abnormality. Fifty-nine patients had presented to the hospital with left lower, 14 with right lower and seven with bilateral lower quadrant pain, and seven subjects complained of left upper quadrant pain. The diagnosis was established preoperatively in 49 patients, intraoperatively in 19, and during the postoperative period in five; 14 patients were aware of having this anomaly. The data of eight patients were not unavailable. Eleven patients underwent laparoscopic appendectomy, which was combined with cholecystectomy in two cases. Histopathological examination of the appendix specimens revealed adenocarcinoma in only two of 95 patients. The diagnosis of left lower quadrant pain is based on well-established clinical symptoms, physical examination and physician's experience.

  10. Cardiac pacemaker

    International Nuclear Information System (INIS)

    Kolenik, S.A.

    1976-01-01

    The construction of a cardiac pacemaker is described which is characterized by particularly small dimensions, small weight and long life duration. The weight is under 100g, the specific weight under 1.7. Mass inertia forces which occur through acceleration and retardation processes, thus remain below the threshold values, above which one would have to reckon with considerable damaging of the surrounding body tissue. The maintaining of small size and slight weight is achieved by using an oscillator on COSMOS basis, where by considerably lower energy consumption, amongst others the lifetimes of the batteries used - a lithium anode with thionyl chloride electrolyte - is extended to over 5 years. The reliability can be increased by the use of 2 or more batteries. The designed dimension are 20x60x60 mm 3 . (ORU/LH) [de

  11. Possible better long-term survival in left versus right-sided colon cancer - a systematic review

    DEFF Research Database (Denmark)

    Hansen, Iben Onsberg; Jess, Per

    2012-01-01

    Colon cancer is one of the most frequent types of cancer in Denmark and the western world. Recent studies indicate that there are differences between right- and left-sided colon cancer with regard to epidemiology, clinical manifestation, pathology and prognosis. The present systematic literature...

  12. Leadless Cardiac Pacemakers: Current status of a modern approach in pacing

    Directory of Open Access Journals (Sweden)

    Skevos Sideris

    2017-11-01

    Full Text Available Since the first transvenous pacemaker implantation, which took place 50 years ago, important progress has been achieved in pacing technology. Consequently, at present, more than 700,000 pacemakers are implanted annually worldwide. However, conventional pacemakers' implantation has a non-negligible risk of periprocedural and long-term complications associated with the transvenous leads and pacemaker pocket. Recently, leadless pacing systems have emerged as a therapeutic alternative to conventional pacing systems that provide therapy for patients with bradyarrhythmias, while eliminating potential transvenous lead- and pacemaker pocket-related complications. Initial studies have demonstrated favorable efficacy and safety of currently developed leadless pacing systems, compared to transvenous pacemakers. In the present paper, we review the current evidence and highlight the advantages and disadvantages of this novel technology. New technological advances may allow the next generation of leadless pacemakers to further expand, thereby offering a wireless cardiac pacing in future. Keywords: cardiac pacing, pacemaker, leadless pacemaker, bradycardia

  13. Transcatheter leadless pacemaker implantation in a patient with a transvenous dual-chamber pacemaker already in place.

    Science.gov (United States)

    Karjalainen, Pasi P; Nammas, Wail; Paana, Tuomas

    2016-01-01

    An 83-year-old lady had a DDDR pacemaker inserted in 1997 for symptomatic atrioventricular block. She underwent battery replacement in 2008. In 2010, she developed atrial fibrillation; the pacemaker was switched to VVIR mode. During the last 2years, ventricular lead threshold increased progressively. In December 2015, she presented for elective battery replacement. After successful battery replacement, the ventricular lead threshold remained high; therefore, we implanted a leadless transcatheter pacemaker, via femoral vein access, using a dedicated catheter delivery system. Electrical measurements at this stage revealed a pacing threshold of 0.28V at 0.24msec, and an impedance of 650Ω. Copyright © 2016 Elsevier Inc. All rights reserved.

  14. Cardiovascular side effects of iopamidol, metrizamide, and amidotrizoate after intracoronary and left ventricular injections

    Energy Technology Data Exchange (ETDEWEB)

    Schraeder, R.; Wolpers, H.G.; Hoeft, A.; Korb, H.; Hellige, G.

    1984-05-01

    The acute cardiovascular side effects of iopamidol, metrizamide, and amidotrizoate during coronary arteriography and left ventriculography have been compared in anesthetized closed-chest dogs. The influence of these contrast media on hemodynamics, coronary circulation, ECG and physicochemical properties of coronary sinus blood were evaluated. Following intracoronary injection, iopamidol and metrizamide cause positive inotropism while amidotrizoate initially induces cardiodepression. All contrast solutions lead to marked increase of myocardial blood flow. Electrocardiographic changes are greater after amidotrizoate. A fall of cationic content in coronary sinus blood is seen after each contrast medium. Amidotrizoate leads to more decrease of potassium and both ionized and total calcium. Increase of osmolality and hydrogen ionic content is greater after amidotrizoate than after iopamidol and metrizamide. Following left ventricular injection, marked peripheral vasodilation, as seen after amidotrizoate, does not occur after low osmolality compounds. There are no significant differences between iopamidol and metrizamide in spite of the slightly higher osmolality of the former. Thus, both contrast materials should be equivalent with respect to their acute cardiovascular side effects under clinical conditions. Advantages of low osmolality radiological contrast media are discussed.

  15. Pacemaker implantation complication rates in elderly and young patients

    Directory of Open Access Journals (Sweden)

    Özcan KS

    2013-08-01

    Full Text Available Kazim Serhan Özcan, Damirbek Osmonov, Servet Altay, Cevdet Dönmez, Ersin Yildirim, Ceyhan Türkkan, Baris Güngör, Ahmet Ekmekçi, Ahmet Taha Alper, Kadir Gürkan, İzzet ErdinlerDepartment of Cardiology, Siyami Ersek Cardiovascular and Thoracic Surgery Center, Istanbul, TurkeyAims: To evaluate the complication rate differences between elderly and younger patients who receive a permanent pacemaker implantation.Methods: We reviewed all cases admitted to our institution between January 2008 and June 2009 with symptomatic bradyarrhythmia for whom a permanent pacemaker was implanted. Beginning in June 2009, we prospectively collected data from all patients with the same diagnosis and procedure. The frequency of complications due to the pacemaker implantation procedure was evaluated and compared between young (<70 years old and elderly (≥70 years old patients.Results: Among 574 patients with a permanent pacemaker, 259 patients (45.1% were below and 315 patients (54.9% were above or at 70 years of age. There were 240 (92.7% and 19 (7.3% dual-chamber pacemaker (DDD and single-chamber pacemaker (VVI implanted patients in the younger group, and 291 (76.8% and 73 (23.2% DDD and VVI pacemaker implanted patients in the elderly group, respectively. The complication rate was 39 (15.1% out of 259 young patients and 24 (7.6% out of 315 elderly patients. Postprocedural complications were statistically lower in the elderly patients than in younger patients (P = 0.005.Conclusion: A pacemaker implantation performed by an experienced operator is a safe procedure for patients of advanced age. The patients who are above 70 years old may have less complication rates than the younger patients.Keywords: complications of pacemaker implantation, elderly patients, permanent pacemaker

  16. VMAT techniques for lymph node-positive left sided breast cancer

    Energy Technology Data Exchange (ETDEWEB)

    Pasler, Marlies; Lutterbach, Johannes; Bjoernsgard, Mari; Reichmann, Ursula; Bartelt, Susanne [Lake Constance Radiation Oncology Center Singen, Friedrichshafen (Germany); Georg, Dietmar [Medical Univ. Vienna/AKH Vienna (Austria). Dept. of Radiooncology; Medical Univ. Vienna (Austria). Christian Doppler Lab. for Medical Radiation Research for Radiation Oncology

    2015-09-01

    To investigate the plan quality of two different volumetric modulated arc therapy (VMAT) techniques for lymph node-positive left-sided breast cancer. Two VMAT plans were generated for 10 lymph node-positive left-sided breast cancer patients: one plan using one single segment of a full rotation, typically an arc segment of 230 (1s-VMAT); and a second plan consisting of 2 small tangential arc segments of about 50 (2s-VMAT). For plan comparison, various dose and dose volume metrics (D{sub mean}, D{sub 98%}, D{sub 2%} for target volumes, D{sub 2%}, D{sub mean} and V{sub x%} for organs at risk (OAR)) were evaluated. Both techniques fulfilled both clinical target dose and OAR goals. 1s-VMAT achieved a slightly better homogeneity and better target coverage (D{sub 2%} = 54.2 ± 0.7 Gy, D{sub 98%} = 30.3 ± 1.8 Gy) compared to 2s-VMAT (D{sub 2%} = 55.0 ± 1.1 Gy, D{sub 98%} = 29.9 ± 1.7 Gy). For geometrical reasons, OAR sparing was noticeable but not significant better using 2s-VMAT, particularly heart and contralateral breast. The heart received a mean dose of 4.4 ± 0.8 Gy using 1s-VMAT and 3.3 ± 1.0 Gy using 2s-VMAT; the contralateral breast received 1.5 ± 0.3 Gy and 0.9 ± 0.3 Gy, respectively. A VMAT technique based on two small tangential arc segments enables improved OAR sparing; the differences between the two techniques in target coverage and homogeneity are minor. Patient age and -anatomy must be considered for each individual case when deciding which technique to be used.

  17. 21 CFR 870.5550 - External transcutaneous cardiac pacemaker (noninvasive).

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false External transcutaneous cardiac pacemaker... § 870.5550 External transcutaneous cardiac pacemaker (noninvasive). (a) Identification. An external transcutaneous cardiac pacemaker (noninvasive) is a device used to supply a periodic electrical pulse intended to...

  18. 21 CFR 870.3710 - Pacemaker repair or replacement material.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Pacemaker repair or replacement material. 870.3710... (CONTINUED) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Prosthetic Devices § 870.3710 Pacemaker repair or replacement material. (a) Identification. A pacemaker repair or replacement material is an...

  19. 21 CFR 870.1750 - External programmable pacemaker pulse generator.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false External programmable pacemaker pulse generator. 870.1750 Section 870.1750 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN... External programmable pacemaker pulse generator. (a) Identification. An external programmable pacemaker...

  20. Low permanent pacemaker rates following Lotus device implantation for transcatheter aortic valve replacement due to modified implantation protocol.

    Science.gov (United States)

    Krackhardt, Florian; Kherad, Behrouz; Krisper, Maximilian; Pieske, Burkert; Laule, Michael; Tschöpe, Carsten

    2017-01-01

    Conduction disturbances requiring permanent pacemaker implantation following transcatheter aortic valve replacement (TAVR) are a common problem. Pacemaker implantation rates after TAVR appear to be higher compared to conventional aortic valve replacement. The aim of this study was to analyze whether a high annulus implantation conveys the benefit of a decreased rate of permanent pacemaker implantation while being safe and successful according to Valve Academic Research Consortium 2 (VARC2)-criteria. A total of 23 patients with symptomatic severe aortic valve stenosis, an aortic annulus of 19-27 mm and at high risk for surgery were treated with the Lotus valve. In all patients the valve was implanted in a high annulus position via femoral access. The primary device performance endpoint was VARC2-defined device success after 30 days and the primary safety endpoint was the need for permanent pacemaker implantation. The mean age was 73.23 ± 7.65 years, 46% were female, 38% were New York Heart Association class III/IV at baseline. Thirty-day follow-up data were available for all patients. The VARC2-defined device success rate after 30 days was 22/23 (96%). 2/21 (10%) patients required a newly implanted pacemaker due to 3rd degree atrioventricular block. 25% of the patients developed a new left bundle branch block after valvuloplasty or device implantation. 21 of the 23 patients (96%) had no other signs of conduction disturbances after 30 days. The approach of the modified implantation technique of Lotus TAVR device was safe and effective. The incidence of need for a permanent pacemaker following TAVR could be significantly reduced due to adopted implantation protocol.

  1. Radioisotope-powered cardiac pacemaker program. Clinical studies of the nuclear pacemaker model NU-5. Final report

    International Nuclear Information System (INIS)

    1980-06-01

    Beginning in February, 1970, the Nuclear Materials and Equipment Corporation (NUMEC) undertook a program to design, develop and manufacture a radioisotope powered cardiac pacemaker system. The scope of technical work was specified to be: establish system, component, and process cost reduction goals using the prototype Radioisotope Powered Cardiac Pacemaker (RCP) design and develop production techniques to achieve these cost reduction objectives; fabricate radioisotope powered fueled prototype cardiac pacemakers (RCP's) on a pilot production basis; conduct liaison with a Government-designated fueling facility for purposes of defining fueling requirements, fabrication and encapsulation procedures, safety design criteria and quality control and inspection requirements; develop and implement Quality Assurance and Reliability Programs; conduct performance, acceptance, lifetime and reliability tests of fueled RCP's in the laboratory; conduct liaison with the National Institutes of Health and with Government specified medical research institutions selected for the purpose of undertaking clinical evaluation of the RCP in humans; monitor and evaluate, on a continuing basis, all test data; and perform necessary safety analyses and tests. Pacemaker designs were developed and quality assurance and manufacturing procedures established. Prototype pacemakers were fabricated. A total of 126 radioisotope powered units were implanted and have been followed clinically for approximately seven years. Four (4) of these units have failed. Eighty-three (83) units remain implanted and satisfactorily operational. An overall failure rate of less than the target 0.15% per month has been achieved

  2. Comparison of right and left side heart functions in patients with thalassemia major, patients with thalassemia intermedia, and control group.

    Science.gov (United States)

    Noori, Noormohammad; Mohamadi, Mehdi; Keshavarz, Kambiz; Alavi, Seyed Mostafa; Mahjoubifard, Maziar; Mirmesdagh, Yalda

    2013-01-01

    Heart disease is the main cause of mortality and morbidity in patients with beta thalassemia, rendering its early diagnosis vital. We studied and compared echocardiographic findings in patients with beta thalassemia major, patients with beta thalassemia intermedia, and a control group. Eighty asymptomatic patients with thalassemia major and 22 asymptomatic cases with thalassemia intermedia (8-25 years old) were selected from those referred to Ali Asghar Hospital (Zahedan-Iran) between June 2008 and June 2009. Additionally, 80 healthy individuals within the same age and sex groups were used as controls. All the individuals underwent echocardiography, the data of which were analyzed with the Student t-test. The mean value of the pre-ejection period/ejection time ratio of the left ventricle during systole, the diameter of the posterior wall of the left ventricle during diastole, the left and right isovolumic relaxation times, and the right myocardial performance index in the patients with beta thalassemia major and intermedia increased significantly compared to those of the controls, but the other parameters were similar between the two patient groups. The mean values of the left and right pre-ejection periods, left ventricular end systolic dimension, and left isovolumic contraction time in the patients with thalassemia intermedia increased significantly compared to those of the controls. In the left side, myocardial performance index, left ventricular mass index, isovolumic contraction time, and deceleration time exhibited significant changes between the patients with thalassemia major and those with thalassemia intermedia, whereas all the echocardiographic parameters of the right side were similar between these two groups. The results showed that the systolic and diastolic functions of the right and left sides of the heart would be impaired in patients with thalassemia major and thalassemia intermedia. Consequently, serial echocardiography is suggested in

  3. Cardiovascular patients’ experiences of living with pacemaker: Qualitative study

    Directory of Open Access Journals (Sweden)

    Morteza Ghojazadeh

    2015-09-01

    Full Text Available BACKGROUND: A pacemaker implantation is considered major life event for cardiovascular patients, so they will probably have very interesting experiences of living with this device. The aim of this study was to explore the experiences of cardiovascular patients living with the pacemaker. METHODS: In this qualitative study, 27 patients were chosen through purposive sampling to achieve data saturation, and their experiences were examined using semi-structured interviews. The patients’ statements were recorded with their consent and analyzed using content analysis method. RESULTS: Participants’ experiences included three main themes: “Problems and limitations,” “feeling and dealing with pacemaker”, and “sources of comfort” and 10 sub-themes including: physical problems, financial problems, social problems, the first encounter, the feeling of living with the pacemaker, how to cope with pacemaker, satisfaction with pacemaker, good family support, hospital and hospital staff performance, and role of religious beliefs. CONCLUSION: Planning to solve social problems, identifying and changing feelings of patients using pacemakers, reinforcing the resources of comfort especially family support seem to be necessary steps for improving quality of life and impact of using pacemaker

  4. A case of left-sided unilateral spatial neglect owing to the cerebral infarction in the distribution of the right posterior cerebral artery

    International Nuclear Information System (INIS)

    Odagaki, Yuji; Oka, Iori; Kon, Yutaka; Asano, Yutaka

    1985-01-01

    We report a case of left-sided unilateral spatial neglect (USN) induced by the cerebral infarction in the distribution of right posterior cerebral artery (PCA). A 69-year-old, right-handed man, who had had a sudden onset of left hemiparesis in August 1983, was admitted to our hospital on January 16, 1984, because of nocturnal delirium. He became alert a few days after admission, but was euphoric and sometimes irritable. Neurologic examination disclosed left homonymous hemianopsia, dysarthria, left central facial weakness, spastic left hemiparesis, hyperactive reflexes on the left with no Babinski sign, left hemisensory loss, and left thalamic pain. On neuropsychologic examination it was revealed that he had a tendency to neglect the left half of his extrapersonal space. When asked to locate cities on a blank map of Japan, he located most of them not only on the right side of the map but also incorrectly. He also had a severe acalculia. There was gradual improvement in these neuropsychologic symptoms. CT demonstrated an area of decreased density in the territory of the right PCA, posterolateral portion of the right thalamus, and the posterior limb of right internal capsule, sparing parietal and temporal lobes. Single-photon emission computed tomography (SPECT) using the Xenon-133 inhalation method showed, however, diminished regional cerebral blood flow (rCBF) in an area larger than the area of infarction demonstrated by CT, including the right parieto-temporo-occipital junctional area, which has been considerd to be responsible for left-sided USN. The authors ascribed the patient's left-sided USN to the lesion of this area that was revealed not morphologically by CT but functionally by SPECT, although the possibility that the lesions of the medial portion of the right occipital lobe and/or subcortical lesions of such areas as the thalamus and the internal capsule more or less influenced the neuropsychologic symptoms could not be excluded. (author)

  5. Radiation protection for pacemakers with radionuclide batteries

    International Nuclear Information System (INIS)

    Stieve, F.E.

    1976-01-01

    Radionuclide batteries ( 147 Pm, 238 Pu) in pacemakers bring risks for the patient and the environment, if the radioactive material cannot be secured for subsequent application or for safe end storage. This is why the expenditure connected with pacemakers is very high. So the application of such pacemakers is only indicated when the patient's expectation of life is presumably higher than 5-10 years, when there are no other diseases besides cardiac dysrhytmia, and when the securing of the radionuclide batteries is guaranteed. (orig.) [de

  6. Analysis of a five year experience of permanent pacemaker ...

    African Journals Online (AJOL)

    Introduction: Permanent pacemaker implantation is available in Nigeria. There is however no national registry or framework for pacemaker data collection. A pacemaker database has been developed in our institution and the results are analyzed in this study. Methods: The study period was between January 2008 and ...

  7. Uniform pacemaker and ICD information system in the Netherlands

    NARCIS (Netherlands)

    Cam, H; Wa, D; Callaos, N; Farsi, D; EshaghianWilner, M; Hanratty, T; Rishe, N

    2003-01-01

    The Central Pacemaker Patient Registry (CPPR) in the Netherlands (founded in 1977) collects information of pacemaker patients from all 110 Dutch hospitals. It contains data of over 98.000 patients, 118.500 pacemakers, 1.950 ICD's and 131.000 leads. Initially data was entered manually. As local

  8. A comparison of brain activity associated with language production in brain tumor patients with left and right sided language laterality.

    Science.gov (United States)

    Jansma, J M; Ramsey, N; Rutten, G J

    2015-12-01

    Language dominance is an important factor for clinical decision making in brain tumor surgery. Functional MRI can provide detailed information about the organization of language in the brain. One often used measure derived from fMRI data is the laterality index (LI). The LI is typically based on the ratio between left and right brain activity in a specific region associated with language. Nearly all fMRI language studies show language-related activity in both hemispheres, and as a result the LI shows a large range of values. The clinical significance of the variation in language laterality as measured with the LI is still under debate. In this study, we tested two hypotheses in relation to the LI, measured in Broca's region, and it's right hemisphere homologue: 1: the level of activity in Broca's and it's right hemisphere homologue is mirrored for subjects with an equal but opposite LI; 2: the whole brain language activation pattern differs between subjects with an equal but opposite LI. One hundred sixty-three glioma and meningioma patients performed a verb generation task as part of a standard clinical protocol. We calculated the LI in the pars orbitalis, pars triangularis and pars opercularis of the left inferior frontal gyrus, referred to as Broca's region from here on. In our database, 21 patients showed right lateralized activity, with a moderate average level (-0.32). A second group of 21 patients was selected from the remaining group, for equal but opposite LI (0.32). We compared the level and distribution of activity associated with language production in the left and right hemisphere in these two groups. Patients with left sided laterality showed a significantly higher level of activity in Broca's region than the patients with right sided laterality. However, both groups showed no difference in level of activity in Broca's homologue region in the right hemisphere. Also, we did not see any difference in the pattern of activity between patients with left-sided

  9. Patterns of intrafractional motion and uncertainties of treatment setup reference systems in accelerated partial breast irradiation for right- and left-sided breast cancer.

    Science.gov (United States)

    Yue, Ning J; Goyal, Sharad; Kim, Leonard H; Khan, Atif; Haffty, Bruce G

    2014-01-01

    This study investigated the patterns of intrafractional motion and accuracy of treatment setup strategies in 3-dimensional conformal radiation therapy of accelerated partial breast irradiation (APBI) for right- and left-sided breast cancers. Sixteen right-sided and 17 left-sided breast cancer patients were enrolled in an institutional APBI trial in which gold fiducial markers were strategically sutured to the surgical cavity walls. Daily pre- and postradiation therapy kV imaging were performed and were matched to digitally reconstructed radiographs based on bony anatomy and fiducial markers, respectively, to determine the intrafractional motion. The positioning differences of the laser-tattoo and the bony anatomy-based setups with respect to the marker-based setup (benchmark) were determined to evaluate their accuracy. Statistical differences were found between the right- and left-sided APBI treatments in vector directions of intrafractional motion and treatment setup errors in the reference systems, but less in their overall magnitudes. The directional difference was more pronounced in the lateral direction. It was found that the intrafractional motion and setup reference systems tended to deviate in the right direction for the right-sided breast treatments and in the left direction for the left-sided breast treatments. It appears that the fiducial markers placed in the seroma cavity exhibit side dependent directional intrafractional motion, although additional data may be needed to further validate the conclusion. The bony anatomy-based treatment setup improves the accuracy over laser-tattoo. But it is inadequate to rely on bony anatomy to assess intrafractional target motion in both magnitude and direction. Copyright © 2014 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

  10. Losing the left side of the world: rightward shift in human spatial attention with sleep onset.

    Science.gov (United States)

    Bareham, Corinne A; Manly, Tom; Pustovaya, Olga V; Scott, Sophie K; Bekinschtein, Tristan A

    2014-05-28

    Unilateral brain damage can lead to a striking deficit in awareness of stimuli on one side of space called Spatial Neglect. Patient studies show that neglect of the left is markedly more persistent than of the right and that its severity increases under states of low alertness. There have been suggestions that this alertness-spatial awareness link may be detectable in the general population. Here, healthy human volunteers performed an auditory spatial localisation task whilst transitioning in and out of sleep. We show, using independent electroencephalographic measures, that normal drowsiness is linked with a remarkable unidirectional tendency to mislocate left-sided stimuli to the right. The effect may form a useful healthy model of neglect and help in understanding why leftward inattention is disproportionately persistent after brain injury. The results also cast light on marked changes in conscious experience before full sleep onset.

  11. Aggregatibacter aphrophilus pacemaker endocarditis: a case report.

    Science.gov (United States)

    Patel, Sahil R; Patel, Nishi H; Borah, Amit; Saltzman, Heath

    2014-12-08

    Aggregatibacter bacteria are a rare cause of endocarditis in adults. They are part of a group of organisms known as HACEK--Haemophilus, Aggregatibacter, Cardiobacter, Eikenella, and Kingella. Among these organisms, several Haemophilus species have been reclassified under the genus Aggregatibacter. Very few cases of Aggregatibacter endocarditis in patients with pacemaker devices have been reported. We present here what we believe to be the first case of Aggregatibacter aphrophilus pacemaker endocarditis. A 62-year-old African American male with a medical history significant for dual-chamber pacemaker placement in 1996 for complete heart block with subsequent lead manipulation in 2007, presented to his primary care doctor with fever, chills, night sweats, fatigue, and ten-pound weight loss over a four-month period. Physical examination revealed a new murmur and jugular venous distension which prompted initiation of antibiotics for suspicion of endocarditis. Both sets of initial blood cultures were positive for A. aphrophilus. Transesophageal echocardiogram revealed vegetations on the tricuspid valve and the right ventricular pacemaker lead (Figure 1). This case highlights the importance of identifying rare causes of endocarditis and recognizing that treatment may not differ from the standard treatment for typical presentations. The patient received intravenous ceftriaxone for his endocarditis for a total of six weeks. Upon device removal, temporary jugular venous pacing wires were placed. After two weeks of antibiotic treatment and no clinical deterioration, a new permanent pacemaker was placed and the patient was discharged home. This is the first case of A. aphrophilus endocarditis in a patient with a permanent pacemaker. Our patient had no obvious risk factors other than poor dentition and a history of repeated pacemaker lead manipulation. This suggests that valvulopathies secondary to repeated lead manipulation can be clinically significant factors in morbidity

  12. A case report of percutaneous endoscopic gastrostomy left-side gastropexy to resolve a recurrent gastric dilatation in a dog previously treated with right-side gastropexy for gastric dilatation volvulus.

    Science.gov (United States)

    Spinella, Giuseppe; Cinti, Filippo; Pietra, Marco; Capitani, Ombretta; Valentini, Simona

    2014-12-01

    A 6-year-old, large-breed, female dog was evaluated for gastric dilatation (GD). The dog was affected by GD volvulus, which had been surgically treated with gastric derotation and right incisional gastropexy. Recurrence of GD appeared 36 hours after surgery. The dilatation was immediately treated with an orogastric probe but still recurred 4 times. Therefore, a left-side gastropexy by percutaneous endoscopic gastrostomy (PEG) was performed to prevent intermittent GD. After PEG tube placement, the patient recovered rapidly without side effects. Several techniques of gastropexy have been described as a prophylactic method for gastric dilatation volvulus, but to the authors' knowledge, this is the first report of left-sided PEG gastropexy performed in a case of canine GD recurrence after an incisional right gastropexy. Copyright © 2015 Elsevier Inc. All rights reserved.

  13. Successful pacing using a batteryless sunlight-powered pacemaker.

    Science.gov (United States)

    Haeberlin, Andreas; Zurbuchen, Adrian; Schaerer, Jakob; Wagner, Joerg; Walpen, Sébastien; Huber, Christoph; Haeberlin, Heinrich; Fuhrer, Juerg; Vogel, Rolf

    2014-10-01

    Today's cardiac pacemakers are powered by batteries with limited energy capacity. As the battery's lifetime ends, the pacemaker needs to be replaced. This surgical re-intervention is costly and bears the risk of complications. Thus, a pacemaker without primary batteries is desirable. The goal of this study was to test whether transcutaneous solar light could power a pacemaker. We used a three-step approach to investigate the feasibility of sunlight-powered cardiac pacing. First, the harvestable power was estimated. Theoretically, a subcutaneously implanted 1 cm(2) solar module may harvest ∼2500 µW from sunlight (3 mm implantation depth). Secondly, ex vivo measurements were performed with solar cells placed under pig skin flaps exposed to a solar simulator and real sunlight. Ex vivo measurements under real sunlight resulted in a median output power of 4941 µW/cm(2) [interquartile range (IQR) 3767-5598 µW/cm(2), median skin flap thickness 3.0 mm (IQR 2.7-3.3 mm)]. The output power strongly depended on implantation depth (ρSpearman = -0.86, P pacemaker powered by a 3.24 cm(2) solar module was implanted in vivo in a pig to measure output power and to pace. In vivo measurements showed a median output power of >3500 µW/cm(2) (skin flap thickness 2.8-3.84 mm). Successful batteryless VVI pacing using a subcutaneously implanted solar module was performed. Based on our results, we estimate that a few minutes of direct sunlight (irradiating an implanted solar module) allow powering a pacemaker for 24 h using a suitable energy storage. Thus, powering a pacemaker by sunlight is feasible and may be an alternative energy supply for tomorrow's pacemakers. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

  14. Familial Screening for Left-Sided Congenital Heart Disease: What Is the Evidence? What Is the Cost?

    Directory of Open Access Journals (Sweden)

    Daniel J. Perry

    2017-12-01

    Full Text Available Since the American Heart Association’s recommendation for familial screening of adults with congenital heart disease for bicuspid aortic valve, similar recommendations for other left-sided heart defects, such as hypoplastic left heart syndrome (HLHS, have been proposed. However, defining at-risk populations for these heart defects based on genetics is less straightforward due to the wide variability of inheritance patterns and non-genetic influences such as environmental and lifestyle factors. We discuss whether there is sufficient evidence to standardize echocardiographic screening for first-degree relatives of children diagnosed with HLHS. Due to variations in the inclusion of cardiac anomalies linked to HLHS and the identification of asymptomatic individuals with cardiac malformations, published studies are open to interpretation. We conclude that familial aggregation of obstructive left-sided congenital heart lesions in families with history of HLHS is not supported and recommend that additional screening should adopt a more conservative definition of what truly constitutes this heart defect. More thorough consideration is needed before embracing familial screening recommendations of families of patients with HLHS, since this could inflict serious costs on healthcare infrastructure and further burden affected families both emotionally and financially.

  15. Pacemaker implantation after catheter ablation for atrial fibrillation.

    Science.gov (United States)

    Deshmukh, Abhishek J; Yao, Xiaoxi; Schilz, Stephanie; Van Houten, Holly; Sangaralingham, Lindsey R; Asirvatham, Samuel J; Friedman, Paul A; Packer, Douglas L; Noseworthy, Peter A

    2016-01-01

    Sinus node dysfunction requiring pacemaker implantation is commonly associated with atrial fibrillation (AF), but may not be clinically apparent until restoration of sinus rhythm with ablation or cardioversion. We sought to determine frequency, time course, and predictors for pacemaker implantation after catheter ablation, and to compare the overall rates to a matched cardioversion cohort. We conducted a retrospective analysis using a large US commercial insurance database and identified 12,158 AF patients who underwent catheter ablation between January 1, 2005 and December 31, 2012. Over an average of 2.4 years of follow-up, 5.6 % of the patients underwent pacemaker implantation. Using the Cox proportional hazards models, we found that risk of risks of pacemaker implantation was associated with older age (50-64 and ≥65 versus pacemaker implantation between ablation patients and propensity score (PS)-matched cardioversion groups (3.5 versus. 4.1 % at 1 year and 8.8 versus 8.3 % at 5 years). Overall, pacemaker implantation occurs in about 1/28 patients within 1 year of catheter ablation. The overall implantation rate decreased between 2005 and 2012. Furthermore, the risk after ablation is similar to cardioversion, suggesting that patients require pacing due to a common underlying electrophysiologic substrate, rather than the ablation itself.

  16. Left-sided approach for cardiac procedure and thoracoplasty in a patient with Marfan syndrome.

    Science.gov (United States)

    Shimamura, J; Kubota, H; Sudo, K

    2012-02-01

    We present a case of a 20-year-old female in whom we successfully performed a simultaneous Bentall procedure and thoracoplasty by initially removing only the left side of the costal cartilages. This modified sternal elevation technique offered chest stability and an excellent surgical view, and the postoperative course was satisfactory. Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

  17. Long-term mortality from cardiac causes after adjuvant hypofractionated vs. conventional radiotherapy for localized left-sided breast cancer.

    Science.gov (United States)

    Chan, Elisa K; Woods, Ryan; Virani, Sean; Speers, Caroline; Wai, Elaine S; Nichol, Alan; McBride, Mary L; Tyldesley, Scott

    2015-01-01

    Ongoing concern remains regarding cardiac injury with hypofractionated whole breast/chest-wall radiotherapy (HF-WBI) compared to conventional radiotherapy (CF-WBI) in left-sided breast cancer patients. The purpose was to determine if cardiac mortality increases with HF-WBI relative to CF-WBI. Between 1990 and 1998, 5334 women with early-stage breast cancer received post-operative radiotherapy to the breast/chest wall alone. A population-based database recorded baseline patient, tumor and treatment factors. Baseline cardiovascular risk factors were identified from hospital administrative records. A propensity-score model balanced risk factors between radiotherapy groups. Cause of death was coded as breast cancer, cardiac or other cause. Cumulative mortality from each cause after radiotherapy was estimated using a competing risk approach. For left-sided cases, median follow-up was 14.2 years. 485 women received CF-WBI, 2221 women received HF-WBI. There was no difference in 15-year mortality from cardiac causes: 4.8% with HF-WBI and 4.2% with CF-WBI (p=0.74), even after propensity-score adjustment (p=0.45). There was no difference in breast cancer mortality or other cause mortality. For right-sided cases, there was no difference in mortality for the three causes of death. At 15-years follow-up, cardiac mortality is not statistically different among left-sided breast cancer patients treated with HF-WBI or CF-WBI. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  18. Long-term mortality from cardiac causes after adjuvant hypofractionated vs. conventional radiotherapy for localized left-sided breast cancer

    International Nuclear Information System (INIS)

    Chan, Elisa K.; Woods, Ryan; Virani, Sean; Speers, Caroline; Wai, Elaine S.; Nichol, Alan; McBride, Mary L.; Tyldesley, Scott

    2015-01-01

    Background and purpose: Ongoing concern remains regarding cardiac injury with hypofractionated whole breast/chest-wall radiotherapy (HF-WBI) compared to conventional radiotherapy (CF-WBI) in left-sided breast cancer patients. The purpose was to determine if cardiac mortality increases with HF-WBI relative to CF-WBI. Materials and methods: Between 1990 and 1998, 5334 women with early-stage breast cancer received post-operative radiotherapy to the breast/chest wall alone. A population-based database recorded baseline patient, tumor and treatment factors. Baseline cardiovascular risk factors were identified from hospital administrative records. A propensity-score model balanced risk factors between radiotherapy groups. Cause of death was coded as breast cancer, cardiac or other cause. Cumulative mortality from each cause after radiotherapy was estimated using a competing risk approach. Results: For left-sided cases, median follow-up was 14.2 years. 485 women received CF-WBI, 2221 women received HF-WBI. There was no difference in 15-year mortality from cardiac causes: 4.8% with HF-WBI and 4.2% with CF-WBI (p = 0.74), even after propensity-score adjustment (p = 0.45). There was no difference in breast cancer mortality or other cause mortality. For right-sided cases, there was no difference in mortality for the three causes of death. Conclusions: At 15-years follow-up, cardiac mortality is not statistically different among left-sided breast cancer patients treated with HF-WBI or CF-WBI

  19. Parallel distribution of sexes within left and right uterine horns in Holstein dairy cows: evidence that the effect of side of pregnancy on sex ratio could be breed-specific in cattle.

    Science.gov (United States)

    Gharagozlou, F; Vojgani, M; Akbarinejad, V; Niasari-Naslaji, A; Hemmati, M; Youssefi, R

    2013-11-30

    Dissimilar distribution of male and female calves within left and right uterine horns has been observed in beef cows. A retrospective study was conducted to investigate the effect of side of pregnancy on secondary sex ratio in Holstein dairy cows. Data associated with sex of calves, side of pregnancy, sire, dam, parity number of dam, AI technician, season and year were retrieved from the database of a Holstein dairy farm. In total, data consisted of 6515 birth records from 3155 dams and 244 sires across years 2001-2010. Data were analyzed using logistic regression. There was no difference in proportion of male and female calves between left (52.9% and 47.1%, respectively) and right (53.2% and 46.8%, respectively) uterine horns (P>0.05). AI technician, year, season and parity of dam did not affect secondary sex ratio (P>0.05). Secondary sex ratio of left and right uterine horns, and consequently, overall secondary sex ratio (53.1%) were skewed toward males as compared with hypothetical secondary sex ratio of 50% (Pcows. Copyright © 2013 Elsevier B.V. All rights reserved.

  20. Firing Patterns and Transitions in Coupled Neurons Controlled by a Pacemaker

    International Nuclear Information System (INIS)

    Mei-Sheng, Li; Qi-Shao, Lu; Li-Xia, Duan; Qing-Yun, Wang

    2008-01-01

    To reveal the dynamics of neuronal networks with pacemakers, the firing patterns and their transitions are investigated in a ring HR neuronal network with gap junctions under the control of a pacemaker. Compared with the situation without pacemaker, the neurons in the network can exhibit various firing patterns as the external current is applied or the coupling strength of pacemaker varies. The results are beneficial for understanding the complex cooperative behaviour of large neural assemblies with pacemaker control

  1. Validation of the Netherlands pacemaker patient registry

    NARCIS (Netherlands)

    Dijk, WA; Kingma, T; Hooijschuur, CAM; Dassen, WRM; Hoorntje, JCA; van Gelder, LM

    1997-01-01

    This paper deals with the validation of the information stored in the Netherlands central pacemaker patient database. At this moment the registry database contains information on more than 70500 patients, 85000 pacemakers and 90000 leads. The validation procedures consisted of an internal

  2. On the Evolution of the Cardiac Pacemaker

    NARCIS (Netherlands)

    Burkhard, Silja; van Eif, Vincent; Garric, Laurence; Christoffels, Vincent M.; Bakkers, Jeroen

    2017-01-01

    The rhythmic contraction of the heart is initiated and controlled by an intrinsic pacemaker system. Cardiac contractions commence at very early embryonic stages and coordination remains crucial for survival. The underlying molecular mechanisms of pacemaker cell development and function are still not

  3. Feasibility of pacemaker therapy after dynamic cardiomyoplasty

    NARCIS (Netherlands)

    Van den Berg, MP; Nagelkerke, D; Brouwer, RMHJ; Mulder, H; De Boer, H; Crijns, HJGM

    1999-01-01

    A 54-year-old man presented with total atrioventricular (AV) block 3 months after dynamic cardiomyoplasty was performed because of heart failure due to idiopathic dilated cardiomyopathy. Though the cardiomyostimulator acted as a back-up pacemaker, a DDDR pacemaker was implanted to optimize

  4. Selective interference with pacemaker activity by electrical dental devices.

    Science.gov (United States)

    Miller, C S; Leonelli, F M; Latham, E

    1998-01-01

    We sought to determine whether electromagnetic interference with cardiac pacemakers occurs during the operation of contemporary electrical dental equipment. Fourteen electrical dental devices were tested in vitro for their ability to interfere with the function of two Medtronics cardiac pacemakers (one a dual-chamber, bipolar Thera 7942 pacemaker, the other a single-chamber, unipolar Minix 8340 pacemaker). Atrial and ventricular pacemaker output and electrocardiographic activity were monitored by means of telemetry with the use of a Medtronics 9760/90 programmer. Atrial and ventricular pacing were inhibited by electromagnetic interference produced by the electrosurgical unit up to a distance of 10 cm, by the ultrasonic bath cleaner up to 30 cm, and by the magnetorestrictive ultrasonic scalers up to 37.5 cm. In contrast, operation of the amalgamator, electric pulp tester, composite curing light, dental handpieces, electric toothbrush, microwave oven, dental chair and light, ENAC ultrasonic instrument, radiography unit, and sonic scaler did not alter pacing rate or rhythm. These results suggest that certain electrosurgical and ultrasonic instruments may produce deleterious effects in medically fragile patients with cardiac pacemakers.

  5. Deviating colostomy construction versus stent placement as bridge to surgery for malignant left-sided colonic obstruction

    NARCIS (Netherlands)

    Amelung, F.J.; Borg, F. ter; Consten, E.C.; Siersema, P.D.; Draaisma, W.A.

    2016-01-01

    BACKGROUND: Acute colonic decompression using a deviating colostomy (DC) or a self-expandable metal stent (SEMS) has been shown to lead to fewer complications and permanent stomas compared to acute resection in elderly patients with malignant left-sided colonic obstruction (LSCO). However, no

  6. Connectivity of Pacemaker Neurons in the Neonatal Rat Superficial Dorsal Horn

    Science.gov (United States)

    Ford, Neil C.; Arbabi, Shahriar; Baccei, Mark L.

    2014-01-01

    Pacemaker neurons with an intrinsic ability to generate rhythmic burst-firing have been characterized in lamina I of the neonatal spinal cord, where they are innervated by high-threshold sensory afferents. However, little is known about the output of these pacemakers, as the neuronal populations which are targeted by pacemaker axons have yet to be identified. The present study combines patch clamp recordings in the intact neonatal rat spinal cord with tract-tracing to demonstrate that lamina I pacemaker neurons contact multiple spinal motor pathways during early life. Retrograde labeling of premotor interneurons with the trans-synaptic virus PRV-152 revealed the presence of burst-firing in PRV-infected lamina I neurons, thereby confirming that pacemakers are synaptically coupled to motor networks in the spinal ventral horn. Notably, two classes of pacemakers could be distinguished in lamina I based on cell size and the pattern of their axonal projections. While small pacemaker neurons possessed ramified axons which contacted ipsilateral motor circuits, large pacemaker neurons had unbranched axons which crossed the midline and ascended rostrally in the contralateral white matter. Recordings from identified spino-parabrachial and spino-PAG neurons indicated the presence of pacemaker activity within neonatal lamina I projection neurons. Overall, these results show that lamina I pacemakers are positioned to regulate both the level of activity in developing motor circuits as well as the ascending flow of nociceptive information to the brain, thus highlighting a potential role for pacemaker activity in the maturation of pain and sensorimotor networks in the CNS. PMID:25380417

  7. Right- and left-sided colon cancer - clinical and pathological differences of the disease entity in one organ.

    Science.gov (United States)

    Mik, Michal; Berut, Maciej; Dziki, Lukasz; Trzcinski, Radzislaw; Dziki, Adam

    2017-02-01

    Some researchers suggest that cancers located in the right vs. the left side of the colon are different and they can be regarded as distinct disease entities. The aim of this study was to analyze differences in clinical, epidemiological and pathological features of patients with right-sided (RCC) and left-sided (LCC) colon cancer. One thousand two hundred and twenty-four patients were operated on due to colorectal cancer. A group of 477 patients (254 women, mean age 65.5 ±11 for the whole group) with colon cancer was included (212 RCC vs. 265 LCC). Right colon cancer patients were older (67.8 ±11.3 vs. 63.2 ±11.2; p = 0.0087). Left colon cancer patients underwent surgery for urgent indications more often (17.0% vs. 8.5%; p = 0006). Tumor diameter was greater in the RCC group (55 ±60 mm vs. 38 ±21 mm; p = 0.0003). Total number of removed lymph nodes was higher in the RCC group (11.7 ±6 vs. 8.3 ±5; p = 0.0001). Lymph node ratio was higher in the LCC group (0.45 ±0.28 vs. 0.30 ±0.25; p = 0.0063). We found a strong positive correlation between tumor diameter and the number of removed lymph nodes in the LCC group ( r = 0.531). These differences may result from the fact that RCC patients are diagnosed at an older age. The smaller number of removed lymph nodes in LCC patients may result in incorrect staging. It is still necessary to find other biological dissimilarities of adenocarcinoma located on different sides of the colon.

  8. Node-positive left-sided breast cancer. Does VMAT improve treatment plan quality with respect to IMRT?

    Energy Technology Data Exchange (ETDEWEB)

    Pasler, M.; Bartelt, S.; Lutterbach, J. [Lake Constance Radiation Oncology Center Singen, Friedrichshafen (Germany); Georg, D. [Medical University Vienna/AKH Wien, Vienna (Austria). Dept. of Radiooncology; Medical University Vienna (Austria). Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology

    2013-05-15

    Purpose: The aim of the present work was to explore plan quality and dosimetric accuracy of intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) for lymph node-positive left-sided breast cancer. Methods: VMAT and IMRT plans were generated with the Pinnacle{sup 3} V9.0 treatment planning system for 10 lymph node-positive left-sided breast cancer patients. VMAT plans were created using a single arc and IMRT was performed with 4 beams using 6, 10, and 15 MV photon energy, respectively. Plans were evaluated both manually and automatically using ArtiView trademark. Dosimetric plan verification was performed with a 2D ionization chamber array placed in a full scatter phantom. Results: Photon energy had no significant influence on plan quality for both VMAT and IMRT. Large variability in low doses to the heart was found due to patient anatomy (range V{sub 5} {sub Gy} 26.5-95 %). Slightly more normal tissue dose was found for VMAT (e.g., V{sub Tissue30%} = 22 %) than in IMRT (V{sub Tissue30%} = 18 %). The manual and ArtiView trademark plan evaluation coincided very accurately for most dose metrics (difference < 1 %). In VMAT, 96.7 % of detector points passed the 3 %/3 mm gamma criterion; marginally better accuracy was found in IMRT (98.3 %). Conclusion: VMAT for node-positive left-sided breast cancer retains target homogeneity and coverage when compared to IMRT and allows maximum doses to organs at risk to be reduced. ArtiView trademark enables fast and accurate plan evaluation. (orig.)

  9. Development of the cardiac pacemaker

    Science.gov (United States)

    Liang, Xingqun; Evans, Sylvia M.

    2017-01-01

    The sinoatrial node (SAN) is the dominant pacemaker of the heart. Abnormalities in SAN formation and function can cause sinus arrhythmia, including sick sinus syndrome and sudden death. A better understanding of genes and signaling pathways that regulate SAN development and function is essential to develop more effective treatment to sinus arrhythmia, including biological pacemakers. In this review, we briefly summarize the key processes of SAN morphogenesis during development, and focus on the transcriptional network that drives SAN development. PMID:27770149

  10. Fifty years of pacemaker advancements.

    Science.gov (United States)

    Steinhaus, David

    2008-12-01

    A 1957 power blackout in Minnesota prompted C. Walton Lillehei, MD, a pioneer in open heart surgery, to ask Earl Bakken, the co-founder of Medtronic, Inc., to create a battery-operated pacemaker for pediatric patients. That conversation led to the development of the first external battery-operated pacemaker. That first bulky device is far removed from the tiny implantable computers available to heart patients today. Now, the size of two silver dollars stacked on top of one another, a pacemaker is prescribed for a person whose heart beats too slowly or pauses irregularly. Slightly larger devices have more recently evolved from pacing and regulating the heartbeat to being able to provide therapeutic high voltage shocks when needed to stop runaway fast heart rates, recording heart activity, and other physiologic functions, even resynchronizing the heart's chambers-all while providing information on the patient's condition and device performance to the doctor remotely or in the office.

  11. Electron-beam CT coronary angiography in the patients with high heart rate arrhythmia or pacemaker

    International Nuclear Information System (INIS)

    Dong Zhi; Zhu Jiemin; Liu Zhe; Liu Junbo; Li Youjie; Qi Ji

    2006-01-01

    Objective: To report the clinical applicability of coronary angiography for patients with high heart rate, arrhythmia or cardiac pacing using the new-generation of electron-beam CT (e-Speed). Methods: EBCT (GE e-Speed) coronary angiography was performed in 36 eases (male 27, female 9, mean age 58), including the heart rate more than 90 bpm in 20 patients, frequent ectopic beats in 11 cases, implantation of cardiac pacemaker in 4 patients and the unacceptable MSCT image quality due to variability of interscan heart rate (from 82 bpm to 104 bpm) in 1 case. After volume data set was acquired using spiral mode with prospective ECG-gating, the reconstructions of MIP, CPR, VR and Cine were performed. The VR quality was evaluated using a five-point scale. Results: The quality of coronary imaging in all of 36 cases were acceptable. The total visualization rate of coronary artery branches was 80.0%. Left main, left anterior artery and right coronary artery were visualized in all patients and in 94.3% of all cases circumflex artery were visible. Conclusion: EBCT (e-Speed) is applicable in noninvasive coronary angiography for patients with high heart rate, arrhythmia or implanted cardiac pacemaker', and this examination can obtain satisfied diagnosis. (authors)

  12. Novel experience of laser-assisted 'inside-out' central venous access in a patient with bilateral subclavian vein occlusion requiring pacemaker implantation.

    Science.gov (United States)

    Aye, Thandar; Phan, Thanh Trung; Muir, Douglas Findlay; Linker, Nicholas John; Hartley, Richard; Turley, Andrew John

    2017-10-01

    This new laser facilitated 'inside-out' technique was used for transvenous pacemaker insertion in a pacemaker-dependent patient with bilateral subclavian occlusion and a failed epicardial system who is not suitable for a transfemoral approach. Procedure was undertaken under general anaesthesia with venous access obtained from right femoral vein and left axillary vein. 7F multipurpose catheter was used to enter proximal edge of the occluded segment of subclavian vein via femoral approach, which then supported stiff angioplasty wires and microcatheters to tunnel into the body of occlusion. When encountered with impenetrable resistance, 1.4 mm Excimer laser helped delivery of a Pilot 200 wire, which then progressed towards the distal edge of occlusion. Serial balloon dilatations allowed wire tracked into subintimal plane, advanced towards left clavicle using knuckle wire technique, which was then externalized with blunt dissection from infraclavicular pocket area. It was later changed to Amplatz superstiff wire exiting from both ends to form a rail, which ultimately allowed passage of pacing leads after serial balloon dilatation from clavicular end. Our hybrid 'inside-out' technique permitted transvenous pacemaker insertion without complication and this is, to our knowledge, the first case using laser in this context. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.

  13. Influence of digital and analogue cellular telephones on implanted pacemakers.

    Science.gov (United States)

    Altamura, G; Toscano, S; Gentilucci, G; Ammirati, F; Castro, A; Pandozi, C; Santini, M

    1997-10-01

    The aim of this study was to find out whether digital and analogue cellular 'phones affect patients with pacemakers. The study comprised continuous ECG monitoring of 200 pacemaker patients. During the monitoring certain conditions caused by interference created by the telephone were looked for: temporary or prolonged pacemaker inhibition; a shift to asynchronous mode caused by electromagnetic interference; an increase in ventricular pacing in dual chamber pacemakers, up to the programmed upper rate. The Global System for Mobile Communications system interfered with pacing 97 times in 43 patients (21.5%). During tests on Total Access of Communication System telephones, there were 60 cases of pacing interference in 35 patients (17.5%). There were 131 interference episodes during ringing vs 26 during the on/off phase; (P 4 s) was seen at the pacemaker 'base' sensing value in six patients using the Global system but in only one patient using Total Access. Cellular 'phones may be dangerous for pacemaker patients. However, they can be used safely if patients do not carry the 'phone close to the pacemaker, which is the only place where high risk interference has been observed.

  14. Effect of telecobalt irradiation on the function of implantable pacemaker

    International Nuclear Information System (INIS)

    Toyoda, Michiaki

    1986-01-01

    In patients implanted with a pacemaker, radiotherapy may be chosen as the treatment when malignant tumor is complicated. Therefore, on the assumption that the pacemaker apparatus is exposed to X-ray, 21 lithium cells used for CMOS or TTL circuit were collected before expiration date and irradiated with 60 Co. The pacemaker used were 10 apparatuses of unprogrammed model VVI, 9 apparatuses of programmed model VVI and 2 apparatuses programmed model DVI. Irradiation was done up to 1,000 rads in dividing doses or at 1,000 rads as a single dose. Observations were made for effects on intervals, amplitude and wave shape of stimula to pacemaker, power, sensitivity, refractory phase, and program functions. In conclusion, it was found that pacemaker is sure to be affected considerably for various functions, although no functional arrest occurs, under irradiation up to 1,000 rads of 60 Co in dividing dose. When irradiation at 1,000 rads was given as a single dose, dysfunctions of pacemaker developed in some cases indicating that direct irradiation at high doses is contraindicated for pacemakers using much of LSI-CMOS. (author)

  15. Pacemaker lead fracture associated with weightlifting: a report of two cases.

    Science.gov (United States)

    Deering, J A; Pederson, D N

    1993-12-01

    Two cases of pacemaker lead fracture associated with weight-lifting are presented. This is a rare association which has only recently been described in the literature. In both cases, the pacemaker lead was fractured between the clavicle and the first rib, suggesting crush injury. The chest X-ray, pacemaker telemetry with measurement of lead impedance, and pacemaker reprogramming were all helpful in management.

  16. Do prehospital discharge pacemaker checks provide any additional clinical benefit?

    Science.gov (United States)

    Wheelan, Kevin R; Legge, Darlene M; Sakowski, Brent C; Bruce, Susan S; Roberts, David C; Johnston, L Murphy; Moore, B Jane; Beveridge, Thomas P; Wells, Peter J; Vallabahn, Ravi; Donsky, Michael S; Franklin, Jay O

    2005-08-01

    We performed a retrospective analysis of 250 records of consecutive, newly implanted, pacemaker patients from a single center to determine the rate of postimplant complications and observations discovered before and during the prehospital discharge evaluation. No observations occurred in 246 of 250 patients (98.4%) (1-sided 95% confidence interval 96.4%). Of the 250 patients, 4 had observations that were discovered at the prehospital discharge check and required reprogramming to increase the sensitivity safety margin (3 atrial and 1 ventricular). We documented only 1 complication that was discovered before the predischarge evaluation through telemetry and resulted in an atrial lead revision.

  17. Left atrial accessory appendages, diverticula, and left-sided septal pouch in multi-slice computed tomography. Association with atrial fibrillation and cerebrovascular accidents.

    Science.gov (United States)

    Hołda, Mateusz K; Koziej, Mateusz; Wszołek, Karolina; Pawlik, Wiesław; Krawczyk-Ożóg, Agata; Sorysz, Danuta; Łoboda, Piotr; Kuźma, Katarzyna; Kuniewicz, Marcin; Lelakowski, Jacek; Dudek, Dariusz; Klimek-Piotrowska, Wiesława

    2017-10-01

    The aim of this study is to provide a morphometric description of the left-sided septal pouch (LSSP), left atrial accessory appendages, and diverticula using cardiac multi-slice computed tomography (MSCT) and to compare results between patient subgroups. Two hundred and ninety four patients (42.9% females) with a mean of 69.4±13.1years of age were investigated using MSCT. The presence of the LSSP, left atrial accessory appendages, and diverticula was evaluated. Multiple logistic regression analysis was performed to check whether the presence of additional left atrial structures is associated with increased risk of atrial fibrillation and cerebrovascular accidents. At least one additional left atrial structure was present in 51.7% of patients. A single LSSP, left atrial diverticulum, and accessory appendage were present in 35.7%, 16.0%, and 4.1% of patients, respectively. After adjusting for other risk factors via multiple logistic regression, patients with LSSP are more likely to have atrial fibrillation (OR=2.00, 95% CI=1.14-3.48, p=0.01). The presence of a LSSP was found to be associated with an increased risk of transient ischemic attack using multiple logistic regression analysis after adjustment for other risk factors (OR=3.88, 95% CI=1.10-13.69, p=0.03). In conclusion LSSPs, accessory appendages, and diverticula are highly prevalent anatomic structures within the left atrium, which could be easily identified by MSCT. The presence of LSSP is associated with increased risk for atrial fibrillation and transient ischemic attack. Copyright © 2017 Elsevier B.V. All rights reserved.

  18. Autaptic pacemaker mediated propagation of weak rhythmic activity across small-world neuronal networks

    Science.gov (United States)

    Yilmaz, Ergin; Baysal, Veli; Ozer, Mahmut; Perc, Matjaž

    2016-02-01

    We study the effects of an autapse, which is mathematically described as a self-feedback loop, on the propagation of weak, localized pacemaker activity across a Newman-Watts small-world network consisting of stochastic Hodgkin-Huxley neurons. We consider that only the pacemaker neuron, which is stimulated by a subthreshold periodic signal, has an electrical autapse that is characterized by a coupling strength and a delay time. We focus on the impact of the coupling strength, the network structure, the properties of the weak periodic stimulus, and the properties of the autapse on the transmission of localized pacemaker activity. Obtained results indicate the existence of optimal channel noise intensity for the propagation of the localized rhythm. Under optimal conditions, the autapse can significantly improve the propagation of pacemaker activity, but only for a specific range of the autaptic coupling strength. Moreover, the autaptic delay time has to be equal to the intrinsic oscillation period of the Hodgkin-Huxley neuron or its integer multiples. We analyze the inter-spike interval histogram and show that the autapse enhances or suppresses the propagation of the localized rhythm by increasing or decreasing the phase locking between the spiking of the pacemaker neuron and the weak periodic signal. In particular, when the autaptic delay time is equal to the intrinsic period of oscillations an optimal phase locking takes place, resulting in a dominant time scale of the spiking activity. We also investigate the effects of the network structure and the coupling strength on the propagation of pacemaker activity. We find that there exist an optimal coupling strength and an optimal network structure that together warrant an optimal propagation of the localized rhythm.

  19. An electromagnetic compatibility study of cardiac pacemaker to low frequency interferences

    International Nuclear Information System (INIS)

    Andretzko, J.P.; Hedjiedj, A.; Babouri, A.; Guendouz, L.; Nadi, M.

    2006-01-01

    This paper presents an experimental study of the behaviour of cardiac pacemaker submitted to low frequency electromagnetic interferences. The method used in this study is progressive. It consists in starting from the target (the cardiac pacemaker), identifying and quantifying the disturbances (the source), and then introducing secondary influencing parameters in stepwise fashion. The general problematic consists in checking this immunity in relation with led disruptions and in relation with beaming disruptions. The experimental approach suggests two kind of tests corresponding to the two studied coupling modes. The first one corresponds to a direct applying of the disruptive signal between the pacemaker terminals. The objective of this phase is to determine the characteristics of the signal (amplitude and frequency) which are detected by the pacemaker and which generate modifications of its operation. In the second phase the pacemaker is subjected to a variable low frequency magnetic field. This last interacts with the pacemaker by inductive coupling through the loop formed by the pacemaker and its leads and the surrounding medium. This interaction results in an induced electromotive force between the terminals of the pacemaker which can potentially disturb the operation of this last. The objective of this phase is to characterize the signal (magnetic field) likely to generate these disturbances. Tests are carried out on six single chamber pacemaker and five dual chamber pacemaker. The interfering signal frequencies are 50 Hz, 60 Hz, 10 khz and 25 khz. Tracking and programming of the pacemaker housing is achieved with the telemetry system. In this study, the devices have all been configured in inhibited stimulation (S.S.I. or V.V.I. mode according to the international codification), this configuration being the most widespread. The housing stimulates the basic frequency in the absence o f intrinsic activity, the stimulation can be inhibited in each chamber by a

  20. An electromagnetic compatibility study of cardiac pacemaker to low frequency interferences

    Energy Technology Data Exchange (ETDEWEB)

    Andretzko, J.P.; Hedjiedj, A.; Babouri, A.; Guendouz, L.; Nadi, M. [Nancy-1 Univ. Henri Poincare, Lab. d' Instrumentation Electronique de Nancy, Faculte des Sciences, 54 - Vandoeuvre les Nancy (France)

    2006-07-01

    This paper presents an experimental study of the behaviour of cardiac pacemaker submitted to low frequency electromagnetic interferences. The method used in this study is progressive. It consists in starting from the target (the cardiac pacemaker), identifying and quantifying the disturbances (the source), and then introducing secondary influencing parameters in stepwise fashion. The general problematic consists in checking this immunity in relation with led disruptions and in relation with beaming disruptions. The experimental approach suggests two kind of tests corresponding to the two studied coupling modes. The first one corresponds to a direct applying of the disruptive signal between the pacemaker terminals. The objective of this phase is to determine the characteristics of the signal (amplitude and frequency) which are detected by the pacemaker and which generate modifications of its operation. In the second phase the pacemaker is subjected to a variable low frequency magnetic field. This last interacts with the pacemaker by inductive coupling through the loop formed by the pacemaker and its leads and the surrounding medium. This interaction results in an induced electromotive force between the terminals of the pacemaker which can potentially disturb the operation of this last. The objective of this phase is to characterize the signal (magnetic field) likely to generate these disturbances. Tests are carried out on six single chamber pacemaker and five dual chamber pacemaker. The interfering signal frequencies are 50 Hz, 60 Hz, 10 khz and 25 khz. Tracking and programming of the pacemaker housing is achieved with the telemetry system. In this study, the devices have all been configured in inhibited stimulation (S.S.I. or V.V.I. mode according to the international codification), this configuration being the most widespread. The housing stimulates the basic frequency in the absence o f intrinsic activity, the stimulation can be inhibited in each chamber by a

  1. SU-D-18C-06: Initial Experience with Implementing MRI Safety Guidelines for Patients with Pacemakers - Medical Physicist Perspective

    Energy Technology Data Exchange (ETDEWEB)

    James, J; Place, V; Panda, A [Mayo Clinic, Scottsdale, AZ (United States); Edmonson, H [Mayo Clinic College of Medicine, Rochester, MN (United States); Felmlee, J [Mayo Clinic, Rochester, MN (United States); Pooley, R [Mayo Clinic, Jacksonville, FL (United States)

    2014-06-01

    Purpose: Several institutions have developed MRI guidelines for patients with MR-unsafe or MR-conditional pacemakers. Here we highlight the role of a medical physicist in implementing these guidelines for non-pacemaker dependent patients. Guidelines: Implementing these guidelines requires involvement from several medical specialties and a strong collaboration with the site MRI supervisor to develop a structured workflow. A medical physicist is required to be present during the scan to supervise the MR scanning and to maintain a safety checklist that ensures: 1) uninterrupted patient communication with the technologist, 2) continuous patient physiologic monitoring (e.g. blood pressure and electrocardiography) by a trained nurse, 3) redundant patient vitals monitoring (e.g. pulse oximetry) due to the possibility of in vivo electrocardiography reading fluctuations during image acquisition. A radiologist is strongly recommended to be available to review the images before patients are discharged from the scanner. Pacemaker MRI should be restricted to 1.5T field strength. The MRI sequences should be optimized by the physicist with regards to: a) SAR: limited to <1.5 W/Kg for MR-unsafe pacemakers in normal operating mode, b) RF exposure time: <30 min, c) Coils: use T/R coils but not restricted to such, d) Artifacts: further optimization of sequences whenever image quality is compromised due to the pacemaker. In particular, cardiac, breast and left-shoulder MRIs are most susceptible to these artifacts. Possible strategies to lower the SAR include: a) BW reduction, 2) echo-train-length reduction, 3) increase TR, 4) decrease number of averages, 5) decrease flip angle, 6) reduce slices and/or a combination of all the options. Conclusion: A medical physicist in collaboration with the MR supervisor plays an important role in the supervision/implementation of safe MR scanning of pacemaker patients. Developing and establishing a workflow has enabled our institution to scan over

  2. SU-D-18C-06: Initial Experience with Implementing MRI Safety Guidelines for Patients with Pacemakers - Medical Physicist Perspective

    International Nuclear Information System (INIS)

    James, J; Place, V; Panda, A; Edmonson, H; Felmlee, J; Pooley, R

    2014-01-01

    Purpose: Several institutions have developed MRI guidelines for patients with MR-unsafe or MR-conditional pacemakers. Here we highlight the role of a medical physicist in implementing these guidelines for non-pacemaker dependent patients. Guidelines: Implementing these guidelines requires involvement from several medical specialties and a strong collaboration with the site MRI supervisor to develop a structured workflow. A medical physicist is required to be present during the scan to supervise the MR scanning and to maintain a safety checklist that ensures: 1) uninterrupted patient communication with the technologist, 2) continuous patient physiologic monitoring (e.g. blood pressure and electrocardiography) by a trained nurse, 3) redundant patient vitals monitoring (e.g. pulse oximetry) due to the possibility of in vivo electrocardiography reading fluctuations during image acquisition. A radiologist is strongly recommended to be available to review the images before patients are discharged from the scanner. Pacemaker MRI should be restricted to 1.5T field strength. The MRI sequences should be optimized by the physicist with regards to: a) SAR: limited to <1.5 W/Kg for MR-unsafe pacemakers in normal operating mode, b) RF exposure time: <30 min, c) Coils: use T/R coils but not restricted to such, d) Artifacts: further optimization of sequences whenever image quality is compromised due to the pacemaker. In particular, cardiac, breast and left-shoulder MRIs are most susceptible to these artifacts. Possible strategies to lower the SAR include: a) BW reduction, 2) echo-train-length reduction, 3) increase TR, 4) decrease number of averages, 5) decrease flip angle, 6) reduce slices and/or a combination of all the options. Conclusion: A medical physicist in collaboration with the MR supervisor plays an important role in the supervision/implementation of safe MR scanning of pacemaker patients. Developing and establishing a workflow has enabled our institution to scan over

  3. Low pacemaker incidence with continuous-sutured valves: a retrospective analysis.

    Science.gov (United States)

    Niclauss, Lars; Delay, Dominique; Pfister, Raymond; Colombier, Sebastien; Kirsch, Matthias; Prêtre, René

    2017-06-01

    Background Permanent pacemaker implantation after surgical aortic valve replacement depends on patient selection and risk factors for conduction disorders. We aimed to identify risk criteria and obtain a selected group comparable to patients assigned to transcatheter aortic valve implantation. Methods Isolated sutured aortic valve replacements in 994 patients treated from 2007 to 2015 were reviewed. Demographics, hospital stay, preexisting conduction disorders, surgical technique, and etiology in patients with and without permanent pacemaker implantation were compared. Reported outcomes after transcatheter aortic valve implantation were compared with those of a subgroup including only degenerative valve disease and first redo. Results The incidence of permanent pacemaker implantation was 2.9%. Longer hospital stay ( p = 0.01), preexisting rhythm disorders ( p pacemaker implantation. Although prostheses were sutured with continuous monofilament in the majority of cases (86%), interrupted pledgetted sutures were used more often in the pacemaker group ( p = 0.002). In the subgroup analysis, the incidence of permanent pacemaker implantation was 2%; preexisting rhythm disorders and the suture technique were still major risk factors. Conclusion Permanent pacemaker implantation depends on etiology, preexisting rhythm disorders, and suture technique, and the 2% incidence compares favorably with the reported 5- to 10-fold higher incidence after transcatheter aortic valve implantation. Cost analysis should take this into account. Often dismissed as minor complication, permanent pacemaker implantation increases the risks of endocarditis, impaired myocardial recovery, and higher mortality if associated with prosthesis regurgitation.

  4. Case of pacemaker pocket infection caused by Finegoldia magna.

    Science.gov (United States)

    Hosseini Dehkordi, Seyed Hamed; Osorio, Georgina

    2017-10-01

    Finegoldia magna (formerly called Peptostreptococcus magnus) is a Gram-positive anaerobic coccus which is increasingly recognized as an opportunistic pathogen. We present a case of F. magna associated non-valvular cardiovascular device-related infection in an 83 year-old male who received a permanent pacemaker for sick sinus syndrome seven weeks prior to his presentation. Five weeks after the implantation, the pacemaker and leads were explanted because of clinical evidence of pacemaker pocket infection. He was initially treated with sulfamethoxazole-trimethoprim based on the Gram stain results from the removed pacemaker. However, two weeks later, he was readmitted with sepsis and was successfully treated with ampicillin-sulbactam. Culture results from the pacemaker and pocket as well as blood cultures grew F. magna. Clinicians should be aware of the possibility of F. magna infection when initial gram stain results show "gram positive cocci". Copyright © 2017 Elsevier Ltd. All rights reserved.

  5. Effects of irradiation on the components of implantable pacemakers

    CERN Document Server

    Kawamura, S; Kuga, N; Shiba, T; Hirose, T; Fujimoto, H; Toyoshima, T; Hyodo, K; Matoba, M

    2003-01-01

    The purpose of this study was to examine the effects of irradiation on implantable pacemaker components. The pacemaker was divided into three components: lead wire and electrode, battery, and electrical circuit, and each component was irradiated by X-ray and electron beams, respectively. The pacemaker parameters were measured by both telemetry data of the programmer and directly measured data from the output terminal. The following results were obtained. For the lead wire and electrode, there was no effect on the pacemaker function due to irradiation by X-ray and electron beams. In the case of battery irradiation, there was no change in battery voltage or current up to 236 Gy X-ray dose. In the electrical circuit, the pacemaker reverted to the regular beating rate (fixed-rate mode) immediately after the start of X-ray irradiation, and it continued in this mode during irradiation. In patients with their own heartbeat rhythm, changing to the fixed-rate mode may cause dangerous conditions such as ventricular fib...

  6. Growth Patterns of Fetal Lung Volumes in Healthy Fetuses and Fetuses With Isolated Left-Sided Congenital Diaphragmatic Hernia.

    Science.gov (United States)

    Ruano, Rodrigo; Britto, Ingrid Schwach Werneck; Sananes, Nicolas; Lee, Wesley; Sangi-Haghpeykar, Haleh; Deter, Russell L

    2016-06-01

    To evaluate fetal lung growth using 3-dimensional sonography in healthy fetuses and those with congenital diaphragmatic hernia (CDH). Right and total lung volumes were serially evaluated by 3-dimensional sonography in 66 healthy fetuses and 52 fetuses with left-sided CDH between 20 and 37 weeks' menstrual age. Functions fitted to these parameters were compared for 2 groups: (1) healthy versus those with CDH; and (2) fetuses with CHD who survived versus those who died. Fetal right and total lung volumes as well as fetal observed-to-expected right and total lung volume ratios were significantly lower in fetuses with CDH than healthy fetuses (Pvolume ratios did not vary with menstrual age in healthy fetuses or in those with CDH (independent of outcome). Lung volume rates were lower in fetuses with left-sided CDH compared to healthy fetuses, as well as in fetuses with CDH who died compared to those who survived. The observed-to-expected right and total lung volume ratios were relatively constant throughout menstrual age in fetuses with left-sided CDH, suggesting that the origin of their lung growth abnormalities occurred before 20 weeks and did not progress. The observed-to-expected ratios may be useful in predicting the outcome in fetuses with CDH independent of menstrual age. © 2016 by the American Institute of Ultrasound in Medicine.

  7. Pacemaker lead erosion simulating "Loch Ness Monster": conservative management.

    Science.gov (United States)

    Garg, Naveen; Moorthy, Nagaraja

    2012-12-01

    The majority of pacemaker pocket or lead erosions are due to either mechanical erosion by the bulky pulse generator or secondary to pacemaker pocket infection. We describe an unusual case of delayed pacemaker lead erosion causing extrusion of a portion of the pacing lead, with separate entry and exit points, with the gap filled with new skin formation, simulating the "Loch Ness Monster", which was successfully managed conservatively by surgical reinsertion.

  8. Evaluating the pacemaker effect with the pump parameter of gated blood-pool imaging

    International Nuclear Information System (INIS)

    Cheng Muhua

    1995-01-01

    13 normal controls and 27 patients with ventricular pacemaker had undergone planar gated blood-pool imaging in different conditions. Result shows: (1) Pump parameters can successfully reflect therapeutic effect of pacemaker among them EMP is the most valuable parameter for evaluating the cardiac pumping effect. (2) After implantation of the ventricular pacemaker, the LVEF did not increase, but the CO and EMP was significantly increased. (3) Compared with right ventricular demand pacemaker, the rate-responsive ventricular pacemaker give better hemodynamic benefit at exercise condition. (4) Through restrained cardiac pacemaker the functional change was analyzed on or off pace, and monitoring the cardiac function itself after the pacemaker was implanted

  9. Protection of pacemaker wearers: effects on magnetic fields on the operation of implanted cardiac pacemakers

    International Nuclear Information System (INIS)

    Souques, M.; Lambrozo, J.; Frank, R.; Himbert, C.

    2002-01-01

    The aim of this study was to assess the changes in the behavior of cardiac pacemakers exposed to 50 and 60 Hz magnetic fields generated by industrial current and 20 to 50 khz magnetic fields generated by a household in a booming period - the induction cook top - and to study the incidence of these changes in a population of subjects with implanted pacemakers. This will enabled to give patients advices about dealing with electric transport lines and facilities and with induction cook tops and to advise manufacturers about the risks involved

  10. Nuclear-powered pacemaker fuel cladding study

    International Nuclear Information System (INIS)

    Shoup, R.L.

    1976-01-01

    The composite of metals and alloys used in the fabrication of 238 Pu cardiac pacemaker fuel capsules resists the effects of high temperatures, high mechanical forces, and chemical corrosives and provides more than adequate protection to the fuel pellet even from deliberate attempts to dissolve the cladding in inorganic acids. This does not imply that opening a pacemaker fuel capsule by inorganic acids is impossible but that it would not be a wise choice

  11. Calcium Transient and Sodium-Calcium Exchange Current in Human versus Rabbit Sinoatrial Node Pacemaker Cells

    Directory of Open Access Journals (Sweden)

    Arie O. Verkerk

    2013-01-01

    Full Text Available There is an ongoing debate on the mechanism underlying the pacemaker activity of sinoatrial node (SAN cells, focusing on the relative importance of the “membrane clock” and the “Ca2+ clock” in the generation of the small net membrane current that depolarizes the cell towards the action potential threshold. Specifically, the debate centers around the question whether the membrane clock-driven hyperpolarization-activated current, If, which is also known as the “funny current” or “pacemaker current,” or the Ca2+ clock-driven sodium-calcium exchange current, INaCa, is the main contributor to diastolic depolarization. In our contribution to this journal’s “Special Issue on Cardiac Electrophysiology,” we present a numerical reconstruction of If and INaCa in isolated rabbit and human SAN pacemaker cells based on experimental data on action potentials, If, and intracellular calcium concentration ([Ca2+]i that we have acquired from these cells. The human SAN pacemaker cells have a smaller If, a weaker [Ca2+]i transient, and a smaller INaCa than the rabbit cells. However, when compared to the diastolic net membrane current, INaCa is of similar size in human and rabbit SAN pacemaker cells, whereas If is smaller in human than in rabbit cells.

  12. Does high-power computed tomography scanning equipment affect the operation of pacemakers?

    International Nuclear Information System (INIS)

    Yamaji, Satoshi; Imai, Shinobu; Saito, Fumio; Yagi, Hiroshi; Kushiro, Toshio; Uchiyama, Takahisa

    2006-01-01

    Computed tomography (CT) is widely used in clinical practice, but there has not been a detailed report of its effect on the functioning of pacemakers. During CT, ECGs were recorded in 11 patients with pacemakers and the electromagnetic field in the CT room was also measured. The effect of CT on a pacemaker was also investigated in a human body model with and without shielding by rubber or lead. Transient malfunctions of pacemakers during CT occurred in 6 of 11 patients. The model showed that malfunctioning of the pacemaker was induced by CT scanning and this was prevented by lead but not by rubber. The alternating electrical field was 150 V/m on the CT scanning line, which was lower than the level influencing pacemaker functions. The alternating magnetic field was 15μT on the CT scanning line, which was also lower than the level influencing pacemaker functions. Malfunctions of the pacemaker during CT may be caused by diagnostic radiant rays and although they are transient, the possibility of lethal arrhythmia cannot be ignored. (author)

  13. Lung volumes, ventricular function and pulmonary arterial flow in children operated on for left-sided congenital diaphragmatic hernia: long-term results

    International Nuclear Information System (INIS)

    Abolmaali, Nasreddin; Koch, Arne; Goetzelt, Knut; Vogelberg, Christian; Hahn, Gabriele; Fitze, Guido

    2010-01-01

    To compare MRI-based functional pulmonary and cardiac measurements in the long-term follow-up of children operated on for left-sided congenital diaphragmatic hernia (CDH) with age- and body size-matched healthy controls. Twelve children who received immediate postnatal surgery for closure of isolated left-sided CDH were included and received basic medical examinations, pulmonary function testing and echocardiography. MRI included measurement of lung volume, ventricular function assessment and velocity-encoded imaging of the pulmonary arteries and was compared with the data for 12 healthy children matched for age and body size. While patients' clinical test results were not suspicious, comparison between the MRI data for patients and those for healthy controls revealed significant differences. In patients, the volumes of the left lungs were increased and the tidal volume was larger on the right side. While the stroke volumes of both ventricles were reduced, heart rate and ejection fraction were increased. Flow, acceleration time and cross-sectional area of the left pulmonary artery were reduced. Functional MRI detected pulmonary and cardiac findings in the late follow-up of CDH children which may be missed by standard clinical methods and might be relevant for decisions regarding late outcome and treatment. (orig.)

  14. Lung volumes, ventricular function and pulmonary arterial flow in children operated on for left-sided congenital diaphragmatic hernia: long-term results

    Energy Technology Data Exchange (ETDEWEB)

    Abolmaali, Nasreddin; Koch, Arne [Dresden University of Technology, OncoRay - Molecular and Biological Imaging, Medical Faculty Carl Gustav Carus, Dresden (Germany); Goetzelt, Knut; Vogelberg, Christian [University Clinics Carl Gustav Carus, Dresden University of Technology, Clinic and Policlinic for Pediatrics - Pediatric Pulmonology, Dresden (Germany); Hahn, Gabriele [University Clinics Carl Gustav Carus, Dresden University of Technology, Institute and Policlinic for Radiology - Pediatric Radiology, Dresden (Germany); Fitze, Guido [University Clinics Carl Gustav Carus, Dresden University of Technology, Clinic and Policlinic for Pediatric Surgery, Dresden (Germany)

    2010-07-15

    To compare MRI-based functional pulmonary and cardiac measurements in the long-term follow-up of children operated on for left-sided congenital diaphragmatic hernia (CDH) with age- and body size-matched healthy controls. Twelve children who received immediate postnatal surgery for closure of isolated left-sided CDH were included and received basic medical examinations, pulmonary function testing and echocardiography. MRI included measurement of lung volume, ventricular function assessment and velocity-encoded imaging of the pulmonary arteries and was compared with the data for 12 healthy children matched for age and body size. While patients' clinical test results were not suspicious, comparison between the MRI data for patients and those for healthy controls revealed significant differences. In patients, the volumes of the left lungs were increased and the tidal volume was larger on the right side. While the stroke volumes of both ventricles were reduced, heart rate and ejection fraction were increased. Flow, acceleration time and cross-sectional area of the left pulmonary artery were reduced. Functional MRI detected pulmonary and cardiac findings in the late follow-up of CDH children which may be missed by standard clinical methods and might be relevant for decisions regarding late outcome and treatment. (orig.)

  15. Lung volumes, ventricular function and pulmonary arterial flow in children operated on for left-sided congenital diaphragmatic hernia: long-term results.

    Science.gov (United States)

    Abolmaali, Nasreddin; Koch, Arne; Götzelt, Knut; Hahn, Gabriele; Fitze, Guido; Vogelberg, Christian

    2010-07-01

    To compare MRI-based functional pulmonary and cardiac measurements in the long-term follow-up of children operated on for left-sided congenital diaphragmatic hernia (CDH) with age- and body size-matched healthy controls. Twelve children who received immediate postnatal surgery for closure of isolated left-sided CDH were included and received basic medical examinations, pulmonary function testing and echocardiography. MRI included measurement of lung volume, ventricular function assessment and velocity-encoded imaging of the pulmonary arteries and was compared with the data for 12 healthy children matched for age and body size. While patients' clinical test results were not suspicious, comparison between the MRI data for patients and those for healthy controls revealed significant differences. In patients, the volumes of the left lungs were increased and the tidal volume was larger on the right side. While the stroke volumes of both ventricles were reduced, heart rate and ejection fraction were increased. Flow, acceleration time and cross-sectional area of the left pulmonary artery were reduced. Functional MRI detected pulmonary and cardiac findings in the late follow-up of CDH children which may be missed by standard clinical methods and might be relevant for decisions regarding late outcome and treatment.

  16. Radiotherapy with linear accelerator in a patient with permanent pacemaker, methodology and results of dose determination in a case in vivo

    International Nuclear Information System (INIS)

    Diaz A, P.; Dominguez O, X.; Toledo B, V.; Nungaray, H.; Zazueta L, F.

    2010-09-01

    This work presents the methodology used by the radiotherapy department of the Hospital San Javier in the patients handling with pacemaker. By means of dosimetry to the implant area of the pacemaker together with electrocardiography measurements during the radiotherapy sessions, the pacemaker performance is evaluated with the objective of detecting any anomaly to the same caused by electromagnetic interference and/or accumulated dose. It is necessary to follow specific protocols in these patients type that guarantee their security during the exposition to ionizing radiation since a definitive consent that allows to calculate the failure probability of pacemaker in this therapeutic condition does not exist. (Author)

  17. Rare combination of left-sided congenital diaphragmatic hernia and omphalocele.

    Science.gov (United States)

    Chee, Yuet Yee; Wong, Siu Chun Mabel; Wong, Ming Sum Rosanna

    2017-08-07

    We reported a rare case of left-sided posterolateral congenital diaphragmatic hernia (CDH) and omphalocele, which is not associated with chromosomal abnormalities or other syndromes. Omphalocele was detected antenatally (CDH was not detected in antenatal ultrasound). The patient suffered from respiratory failure secondary to severe pulmonary hypertension. As the combination of CDH and omphalocele is rare and with the abdominal content herniating into the omphalocele instead of the thorax, antenatal diagnosis of such condition can be difficult. Unlike other reported cases in the literature, our patient's respiratory condition has been improving with time and is surviving beyond the infancy period. We believe this to be the first such survival case reported in the literature. © BMJ Publishing Group Ltd (unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

  18. Side-Specific Reward Memories in Honeybees

    Science.gov (United States)

    Gil, Mariana; Menzel, Randolf; De Marco, Rodrigo J.

    2009-01-01

    We report a hitherto unknown form of side-specific learning in honeybees. We trained bees individually by coupling gustatory and mechanical stimulation of each antenna with either increasing or decreasing volumes of sucrose solution offered to the animal's proboscis along successive learning trials. Next, we examined their proboscis extension…

  19. Safety of magnetic resonance scanning without monitoring of patients with pacemakers

    DEFF Research Database (Denmark)

    Bertelsen, Litten; Petersen, Helen Høgh; Philbert, Berit Thornvig

    2017-01-01

    AIMS: The objective of this study was to investigate whether it is safe to perform 1.5-Tesla magnetic resonance imaging (MRI) scans in pacemaker (PM) patients without pulse oximetry or electrocardiogram monitoring and with no special specific absorption rate (SAR) or time limits, provided...

  20. Treatment of vasovagal syncope: pacemaker or crossing legs?

    NARCIS (Netherlands)

    van Dijk, N.; Harms, M. P.; Linzer, M.; Wieling, W.

    2000-01-01

    A 50-year-old male patient continued to experience syncope after implantation of a pacemaker. During cardiovascular examination, the patient showed a typical vasovagal response, with normal pacemaker function. Leg crossing, which prohibits the pooling of blood in the legs and abdomen, at the onset

  1. Runaway pacemaker: a forgotten phenomenon?

    Science.gov (United States)

    Ortega, Daniel F; Sammartino, M Victoria; Pellegrino, Graciela M M; Barja, Luis D; Albina, Gaston; Segura, Eliseo V; Balado, Roberto; Laiño, Ruben; Giniger, Alberto G

    2005-11-01

    Runaway is an uncommon pacemaker dysfunction, characterized by fast and erratic spikes at non-physiological rates. This infrequent but potentially lethal failure mode may be related to low battery voltage. Four single chamber pacemaker patients were analyzed (Medtronic Minix ST 8330, Minneapolis, MN, had been implanted in two patients and two CPI Triumph VR 1124, St Paul, MN, in the other two). They had been admitted because of presyncopal episodes. Typical high rate stimuli at 2000 ppm alternating with pacing at 60-65 ppm were recorded in all ECGs. Lead system tests were normal. The pulse generators had to be replaced.

  2. Sacral Neuromodulation in Patients With a Cardiac Pacemaker

    Directory of Open Access Journals (Sweden)

    Abdullah A. Gahzi

    2016-09-01

    Full Text Available The objective of this study was to describe our experience using sacral neuromodulation to treat urinary urgency, frequency, urge incontinence, and chronic urinary retention in patients with cardiac pacemakers. With the increasingly widespread use of InterStim for bladder function restoration, we are seeing more complex patients with multiple comorbidities, including cardiac conditions. Herein, we report 3 cases of individuals with cardiac pacemakers who underwent InterStim implantation to treat urinary conditions. This study is a case series of 3 patients with cardiac pacemakers who underwent sacral neuromodulation to treat refractory voiding dysfunction. The initial patient screening for InterStim therapy involved percutaneous nerve evaluation (PNE, in which a temporary untined lead wire was placed through the S3 foramen. Patients who did not respond to PNE proceeded to a staged implant. All patients in this study had a greater than 50% improvement of their urinary symptoms during the initial trial and underwent placement of the InterStim implantable pulse generator (IPG. Postoperative programming was done under electrocardiogram monitoring by a cardiologist. No interference was observed between the Inter-Stim IPG and the cardiac pacemaker. In this group of patients, sacral neuromodulation in the presence of a cardiac pacemaker appears to have been safe.

  3. Measuring pacemaker dose: A clinical perspective

    Energy Technology Data Exchange (ETDEWEB)

    Studenski, Matthew T., E-mail: matthew.studenski@jeffersonhospital.org [Department of Radiation Oncology at the Jefferson Medical College and Kimmel Cancer Center, Thomas Jefferson University, Philadelphia, PA (United States); Xiao Ying; Harrison, Amy S. [Department of Radiation Oncology at the Jefferson Medical College and Kimmel Cancer Center, Thomas Jefferson University, Philadelphia, PA (United States)

    2012-07-01

    Recently in our clinic, we have seen an increased number of patients presenting with pacemakers and defibrillators. Precautions are taken to develop a treatment plan that minimizes the dose to the pacemaker because of the adverse effects of radiation on the electronics. Here we analyze different dosimeters to determine which is the most accurate in measuring pacemaker or defibrillator dose while at the same time not requiring a significant investment in time to maintain an efficient workflow in the clinic. The dosimeters analyzed here were ion chambers, diodes, metal-oxide-semiconductor field effect transistor (MOSFETs), and optically stimulated luminescence (OSL) dosimeters. A simple phantom was used to quantify the angular and energy dependence of each dosimeter. Next, 8 patients plans were delivered to a Rando phantom with all the dosimeters located where the pacemaker would be, and the measurements were compared with the predicted dose. A cone beam computed tomography (CBCT) image was obtained to determine the dosimeter response in the kilovoltage energy range. In terms of the angular and energy dependence of the dosimeters, the ion chamber and diode were the most stable. For the clinical cases, all the dosimeters match relatively well with the predicted dose, although the ideal dosimeter to use is case dependent. The dosimeters, especially the MOSFETS, tend to be less accurate for the plans, with many lateral beams. Because of their efficiency, we recommend using a MOSFET or a diode to measure the dose. If a discrepancy is observed between the measured and expected dose (especially when the pacemaker to field edge is <10 cm), we recommend analyzing the treatment plan to see whether there are many lateral beams. Follow-up with another dosimeter rather than repeating multiple times with the same type of dosimeter. All dosimeters should be placed after the CBCT has been acquired.

  4. [TRENDS OF PERMANENT PACEMAKER IMPLANTATION IN A SINGLE CENTER OVER A 20-YEAR PERIOD].

    Science.gov (United States)

    Antonelli, Dante; Ilan, Limor Bushar; Freedberg, Nahum A; Feldman, Alexander; Turgeman, Yoav

    2015-05-01

    To review the changes in permanent pacemaker implantation indications, pacing modes and patients' demographics over a 20-year period. We retrospectively retrieved data on patients who underwent first implantation of the pacemaker between 1-1-1991 and 31-12-2010. One thousand and nine (1,009) patients underwent a first pacemaker implantation during that period; 535 were men (53%), their mean age was 74.6±19.5 years; the highest rate of implanted pacemaker was in patients ranging in age from 70-79 years, however there was an increasing number of patients aged over 80 years. The median survival time after initial pacemaker implantation was 8 years. Syncope was the most common symptom (62.5%) and atrioventricular block was the most common electrocardiographic indication (56.4%) leading to pacemaker implantation. There was increased utilization of dual chamber and rate responsive pacemakers over the years. There was no difference regarding mode selection between genders. Pacemaker implantation rates have increased over a 20-year period. Dual chamber replaced most of the single ventricular chamber pacemaker and rate responsive pacemakers became the norm. The data of a small volume center are similar to those reported in pacemaker surveys of high volume pacemaker implantation centers. They confirm adherence to the published guidelines for pacing.

  5. Electromagnetic and radiation environments: effects on pacemakers

    International Nuclear Information System (INIS)

    Mouton, J.; Trochet, R.; Vicrey, J.; Sauvage, M.; Chauvenet, B.; Ostrovski, A.; Leroy, E.; Haug, R.; Dodinot, B.; Joffre, F.

    1999-01-01

    Nowadays, medical care development allows many people to share the benefits of implanted pacemakers (PM). PM can be perturbed and even fall in complete breakdowns in an electromagnetic and radiation environment. A stimuli-dependent patient can thus be seriously in danger. This article presents the effect of ionizing radiation from either a cobalt-60 source or from a linear accelerator (Saturne 43) on 12 pacemakers. It seems that technological progress make electronic circuits more sensitive to the cumulated dose of radiation. This survey shows that pacemakers have great difficulties to sustain ionizing radiation doses that are commonly delivered to patients during therapies. Usually perturbed functioning appears suddenly and means a strong shift of stimuli that might lead to heart failure

  6. Warfarin therapy and incidence of cerebrovascular complications in left-sided native valve endocarditis

    DEFF Research Database (Denmark)

    Snygg-Martin, U; Rasmussen, Rasmus Vedby; Hassager, C

    2011-01-01

    Anticoagulant therapy has been anticipated to increase the risk of cerebrovascular complications (CVC) in native valve endocarditis (NVE). This study investigates the relationship between ongoing oral anticoagulant therapy and the incidence of symptomatic CVC in left-sided NVE. In a prospective...... factors for CVC, while warfarin on admission (aOR 0.26, 95% CI 0.07-0.94), history of congestive heart failure (adjusted OR 0.22, 95% CI 0.1-0.52) and previous endocarditis (aOR 0.1, 95% CI 0.01-0.79) correlated with lower CVC frequency....

  7. Association Between Tangential Beam Treatment Parameters and Cardiac Abnormalities After Definitive Radiation Treatment for Left-Sided Breast Cancer

    International Nuclear Information System (INIS)

    Correa, Candace R.; Das, Indra J.; Litt, Harold I.; Ferrari, Victor; Hwang, W.-T.; Solin, Lawrence J.; Harris, Eleanor E.

    2008-01-01

    Purpose: To examine the association between radiation treatment (RT) parameters, cardiac diagnostic test abnormalities, and clinical cardiovascular diagnoses among patients with left-sided breast cancer after breast conservation treatment with tangential beam RT. Methods and Materials: The medical records of 416 patients treated between 1977 and 1995 with RT for primary left-sided breast cancer were reviewed for myocardial perfusion imaging and echocardiograms. Sixty-two patients (62/416, 15%) underwent these cardiac diagnostic tests for cardiovascular symptoms and were selected for further study. Central lung distance and maximum heart width and length in the treatment field were determined for each patient. Medical records were reviewed for cardiovascular diagnoses and evaluation of cardiac risk factors. Results: At a median of 12 years post-RT the incidence of cardiac diagnostic test abnormalities among symptomatic left-sided irradiated women was significantly higher than the predicted incidence of cardiovascular disease in the patient population, 6/62 (9%) predicted vs. 24/62 (39%) observed, p 0.001. As compared with patients with normal tests, patients with cardiac diagnostic test abnormalities had a larger median central lung distance (2.6 cm vs. 2.2 cm, p = 0.01). Similarly, patients with vs. without congestive heart failure had a larger median central lung distance (2.8 cm vs. 2.3 cm, p = 0.008). Conclusions: Contemporary RT for early breast cancer may be associated with a small, but potentially avoidable, risk of cardiovascular morbidity that is associated with treatment technique

  8. Nuclear-powered pacemaker fuel cladding study

    International Nuclear Information System (INIS)

    Shoup, R.L.

    1976-07-01

    The fabrication of fuel capsules with refractory metal and alloy clads used in nuclear-powered cardiac pacemakers precludes the expedient dissolution of the clad in inorganic acid solutions. An experiment to measure penetration rates of acids on commonly used fuel pellet clads indicated that it is not impossible, but that it would be very difficult to dissolve the multiple cladding. This work was performed because of a suggestion that a 238 PuO 2 -powered pacemaker could be transformed into a terrorism weapon

  9. Opposing nodal and BMP signals regulate left-right asymmetry in the sea urchin larva.

    Directory of Open Access Journals (Sweden)

    Yi-Jyun Luo

    Full Text Available Nodal and BMP signals are important for establishing left-right (LR asymmetry in vertebrates. In sea urchins, Nodal signaling prevents the formation of the rudiment on the right side. However, the opposing pathway to Nodal signaling during LR axis establishment is not clear. Here, we revealed that BMP signaling is activated in the left coelomic pouch, specifically in the veg2 lineage, but not in the small micromeres. By perturbing BMP activities, we demonstrated that BMP signaling is required for activating the expression of the left-sided genes and the formation of the left-sided structures. On the other hand, Nodal signals on the right side inhibit BMP signaling and control LR asymmetric separation and apoptosis of the small micromeres. Our findings show that BMP signaling is the positive signal for left-sided development in sea urchins, suggesting that the opposing roles of Nodal and BMP signals in establishing LR asymmetry are conserved in deuterostomes.

  10. Generic environmental statement on the routine use of plutonium-powered cardiac pacemakers

    International Nuclear Information System (INIS)

    Shoup, R.L.; Robinson, T.W.; O'Donnell, F.R.

    1976-01-01

    The purpose of a continuing program at ORNL is to provide technical assistance to the NRC on writing and editing of the final environmental statement on the routine use of nuclear-powered (primarily 238 Pu) cardiac pacemakers. This environmental statement defines the safety and reliability standards that nuclear-powered pacemakers are required to meet. All aspects of the risks to the patients, the public, and the environment are evaluated both for the routine use of plutonium-powered pacemakers and for postulated accidents involving pacemaker patients. Benefits derived from the use of plutonium-powered units are discussed and weighed against the risks in order to determine whether routine use is justified. Available alternative pacemakers with various performance characteristics are compared with respect to costs and to the needs of pacemaker patients

  11. Global health resource utilization associated with pacemaker complications.

    Science.gov (United States)

    Waweru, Catherine; Steenrod, Anna; Wolff, Claudia; Eggington, Simon; Wright, David Jay; Wyrwich, Kathleen W

    2017-07-01

    To estimate health resource utilization (HRU) associated with the management of pacemaker complications in various healthcare systems. Electrophysiologists (EPs) from four geographical regions (Western Europe, Australia, Japan, and North America) were invited to participate. Survey questions focused on HRU in the management of three chronic pacemaker complications (i.e. pacemaker infections requiring extraction, lead fractures/insulation breaches requiring replacement, and upper extremity deep venous thrombosis [DVT]). Panelists completed a maximum of two web-based surveys (iterative rounds). Mean, median values, and interquartile ranges were calculated and used to establish consensus. Overall, 32 and 29 panelists participated in the first and second rounds of the Delphi panel, respectively. Consensus was reached on treatment and HRU associated with a typical pacemaker implantation and complications. HRU was similar across regions, except for Japan, where panelists reported the longest duration of hospital stay in all scenarios. Infections were the most resource-intensive complications and were characterized by intravenous antibiotics days of 9.6?13.5 days and 21.3?29.2 days for pocket and lead infections respectively; laboratory and diagnostic tests, and system extraction and replacement procedures. DVT, on the other hand, was the least resource intensive complication. The results of the panel represent the views of the respondents who participated and may not be generalizable outside of this panel. The surveys were limited in scope and, therefore, did not include questions on management of acute complications (e.g. hematoma, pneumothorax). The Delphi technique provided a reliable and efficient approach to estimating resource utilization associated with chronic pacemaker complications. Estimates from the Delphi panel can be used to generate costs of pacemaker complications in various regions.

  12. Risk of pacemaker implantation after uneventful successful cavotricuspid isthmus radiofrequency ablation in patients with common atrial flutter.

    Science.gov (United States)

    Rodríguez-Mañero, Moisés; González-Melchor, Layla; Ballesteros, Gabriel; Raposeiras-Roubín, Sergio; García-Seara, Javier; López, Xesús Alberte Fernández; Cambeiro, Cristina González; Alcalde, Oscar; García-Bolao, Ignacio; Martínez-Sande, Luis; González-Juanatey, José Ramón

    2016-01-01

    Little is known about the risk of pacemaker implantation after common atrial flutter ablation in the long-term. We retrospectively reviewed the electrophysiology laboratory database at two Spanish University Hospitals from 1998 to 2012 to identify patients who had undergone successful ablation for cavotricuspid dependent atrial flutter. Cox regression analysis was used to examine the risk of pacemaker implantation. A total of 298 patients were considered eligible for inclusion. The mean age of the enrolled patients was 65.7±11. During 57.7±42.8 months, 30 patients (10.1%) underwent pacemaker implantation. In the stepwise multivariate models only heart rate at the time of the ablation (OR: 0.96; 95% CI: 0.93-0.98; ppacemaker implantation. A heart rate of ≤65 bpm was identified as the optimal cut-off value to predict the need of pacemaker implantation in the follow-up (sensitivity: 79%, specificity: 74%) by ROC curve analyses. This is the first study of an association between the slow conducting common atrial flutter and subsequent risk of pacemaker implantation. In light of these findings, assessing it prior to ablation can be helpful for the risk stratification of sinus node disease or atrioventricular conduction disease requiring a pacemaker implantation in patients with persistent atrial flutter. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  13. Influence of internal current and pacing current on pacemaker longevity.

    Science.gov (United States)

    Schuchert, A; Kuck, K H

    1994-01-01

    The effects of lower pulse amplitude on battery current and pacemaker longevity were studied comparing the new, small-sized VVI pacemaker, Minix 8341, with the former model, Pasys 8329. Battery current was telemetrically measured at 0.8, 1.6, 2.5, and 5.0 V pulse amplitude and 0.05, 0.25, 0.5, and 1.0 msec pulse duration. Internal current was assumed to be equal to the battery current at 0.8 V and 0.05 msec. Pacing current was calculated subtracting internal current from battery current. The Minix pacemaker had a significantly lower battery current because of a lower internal current (Minix: 4.1 +/- 0.1 microA; Pasys: 16.1 +/- 0.1 microA); pacing current of both units was similar. At 0.5 msec pulse duration, the programming from 5.0-2.5 V pulse amplitude resulted in a greater relative reduction of battery current in the newer pacemaker (51% vs 25%). Projected longevity of each pacemaker was 7.9 years at 5.0 V and 0.5 msec. The programming from 5.0-2.5 V extended the projected longevity by 2.3 years (Pasys) and by 7.1 years (Minix). The longevity was negligibly longer after programming to 1.6 V. extension of pacemaker longevity can be achieved with the programming to 2.5 V or less if the connected pacemakers need a low internal current for their circuitry.

  14. Outcome and management of pacemaker-induced superior vena cava syndrome.

    Science.gov (United States)

    Fu, Hai-Xia; Huang, Xin-Miao; Zhong, Li; Osborn, Michael J; Bjarnason, Haraldur; Mulpuru, Siva; Zhao, Xian-Xian; Friedman, Paul A; Cha, Yong-Mei

    2014-11-01

    We aimed to determine the long-term outcomes of percutaneous lead extraction and stent placement in patients with pacemaker-induced superior vena cava (SVC) syndrome. The study retrospectively screened patients who underwent lead extraction followed by central vein stent implantation at Mayo Clinic (Rochester, MN, USA), from January 2005 to December 2012, to identify the patients with pacemaker-induced SVC syndrome. Demographic, clinical, and follow-up characteristics of those patients were collected from electronic medical records. Six cases were identified. The mean (standard deviation) age was 56 (15) years (male, 67%). All patients had permanent dual-chamber pacemakers, with a mean 11-year history of pacemaker placement. The entire device system was explanted in five patients; one patient had a 21-year-old pacemaker lead that could not be removed. Eight stents were implanted in six patients: five patients had one stent, one patient had three. A new pacemaker system was reimplanted through the stented vein in five patients. Technical success was achieved in all patients, without any complication. Symptoms rapidly resolved in all patients after stent deployment. The mean follow-up duration was 48 months (range, 10-100 months). Three patients remained symptom free. Reintervention with percutaneous balloon venoplasty was successful in three patients with symptom recurrence. Percutaneous stent implantation after lead removal followed by reimplantation of leads is a feasible alternative therapy for pacemaker-induced SVC syndrome, although some cases may require repeat intervention. ©2014 Wiley Periodicals, Inc.

  15. Preliminary investigation of orally administered benazepril in horses with left-sided valvular regurgitation.

    Science.gov (United States)

    Afonso, T; Giguère, S; Brown, S A; Barton, M H; Rapoport, G; Barba, M; Dembek, K A; Toribio, R E; Coleman, A E

    2017-10-17

    Despite the paucity of data available, orally administered angiotensin-converting enzyme (ACE) inhibitors are empirically used in horses with valvular regurgitation. Evaluate the echocardiographic and hormonal changes in response to oral benazepril in horses with left-sided valvular regurgitation. Prospective, randomised double-blind, placebo-controlled trial. Horses with mitral valve (MR) and/or aortic valve regurgitation (AR) received oral benazepril (n = 6) at a dosage of 1 mg/kg q 12 h or a placebo (n = 5) for 28 days. Echocardiography was performed before drug administration and after 28 days of treatment. Plasma renin activity, serum ACE activity, angiotensin II concentration, aldosterone concentration and biochemical variables were measured before drug administration and after 7 and 28 days of treatment. Relative to baseline, horses treated with benazepril had statistically significant reduction in left ventricular internal diameter in systole (mean difference between groups = -0.97 cm; 95% CI = -1.5 to -0.43 cm), aortic sinus diameter (-0.31 cm; -0.54 to -0.07 cm), and percentage of the aortic annulus diameter occupied by the base of the AR jet (-17.05%; -31.17 to -2.93%) compared with horses receiving a placebo. In addition, horses treated with benazepril had a significantly greater increase in cardiac output (11.95 L/min; 1.17-22.73 L/min) and fractional shortening (7.59%; 3.3-11.88%) compared with horses receiving a placebo. Despite profound serum ACE inhibition, renin activity and concentrations of angiotensin II and aldosterone were not significantly different between treatment groups or among time points. Very small sample size and short treatment period. Treatment with oral benazepril resulted in statistically significant echocardiographic changes that might indicate reduced cardiac afterload in horses with left-sided valvular regurgitation. Additional studies with a larger sample size will be necessary to determine if administration of benazepril is

  16. Does bipolar pacemaker current activate blood platelets?

    DEFF Research Database (Denmark)

    Gjesdal, Grunde; Hansen, Annebirthe Bo; Brandes, Axel

    2009-01-01

    OBJECTIVE: The aim of this study was to investigate whether bipolar pacemaker current lead can activate blood platelets. The null hypothesis was that 1 minute of electrical stimulation of platelets would not influence their subsequent reactivity to adenosine diphosphate (ADP). BACKGROUND: Both...... platelets and muscle cells contain actin and myosin filaments, and both cells are activated following calcium influx. Muscle cells open their calcium channels and contract when exposed to an electric current. Current through a bipolar pacemaker lead will expose a small volume of blood, including platelets......, to the depolarizing current. Platelet activation may ensue, resulting in aggregation, release reaction, and contraction. In contrast, a unipolar pacemaker system will not depolarize blood, but transmit current directly into the myocardium, and the current afterward passes through other tissues before returning...

  17. [Comparison of clinicopathological features and prognosis between left-sided colon cancer and right-sided colon cancer].

    Science.gov (United States)

    Gao, Xianhua; Yu, Guanyu; Liu, Peng; Hao, Liqiang; Liu, Lianjie; Zhang, Wei

    2017-06-25

    To compare the clinicopathological features and prognosis between left-sided colon cancer (LC) and right-sided colon cancer (RC). Clinicopathological and follow-up data of 2 174 colon carcinoma cases undergoing resection at Shanghai Changhai Hospital of The Second Military Medical University from January 2000 to December 2010 were retrospectively analyzed. Patients with transverse colon cancer, overlapping position, unknown location, recurrent cancer, multiple primary cancer, concomitant malignant tumors, preoperative chemotherapy, local resection, incomplete clinical data and missed follow up were excluded. Finally, a total of 1 036 patients, whose primary tumors were radically removed, were enrolled, with 563 patients in LC group (including carcinoma in cecum, ascending colon and hepatic flexure) and 473 in RC group (including carcinoma in splenic flexure, descending colon and sigmoid colon). The clinicopathological features and survival, including median overall survival, 5-year overall survival rate, tumor specific median overall survival, cancer specific 5-year overall survival rate, were compared between LC and RC groups. Tumor specific overall survival was defined as the period between operation date to the date of death caused by cancer progression. Multivariate Cox regression analysis was used to analyze the influencing factors of survival. Propensity score matching was carried out to balance the clinicopathological factors between the two groups with the SAS 9.3, taking the following parameters into consideration (age, gender, gross appearance, tumor diameter, invasion depth, lymph node metastasis, distant metastasis, TNM stages, differentiation, CEA and CA199-9). Patients in RC group and LC group were matched according to the propensity scores and the clinicopathological characteristics and prognosis of two groups were compared again. No significant differences were identified between the two groups in age, distant metastasis and serum CEA level

  18. 76 FR 64223 - Cardiovascular Devices; Reclassification of External Pacemaker Pulse Generator Devices

    Science.gov (United States)

    2011-10-17

    ... Drug Administration 21 CFR Part 870 Cardiovascular Devices; Reclassification of External Pacemaker... Special Controls Guidance Document: External Pacemaker Pulse Generator; Availability; Proposed Rule and... [Docket No. FDA-2011-N-0650] Cardiovascular Devices; Reclassification of External Pacemaker Pulse...

  19. The cardiac pacemaker patient. Might the pacer be directly irradiated?

    International Nuclear Information System (INIS)

    Tsekos, A.; Momm, F.; Brunner, M.; Guttenberger, R.

    2000-01-01

    According to recommendations listed in a recent review, pacemakers generally should not be irradiated directly. They should be shielded from ionizing radiation or their position changed. Direct irradiation can cause pacemaker failure. There are no clinical signals to indicate a near failure. When controlling pacemaker function, there is often a change in the magnetic frequency before malfunction occurs. If a malfunction results in a partial heart failure only, there would be enough time to implant another pacemaker. As can be seen from a case described here, a policy of 'wait and see' may be adopted even if there are changes in the magnetic frequency. if there is any danger to the patient's life, the feasibility of using an external pacemaker system may be explored, especially in cases where there is a strong need for radiotherapy (severe pain, infiltration of plexus). Risks and benefits should be carefully weighed and discussed with the patient

  20. Testing of Common Electromagnetic Environments for Risk of Interference with Cardiac Pacemaker Function

    Directory of Open Access Journals (Sweden)

    Maria Tiikkaja

    2013-09-01

    Conclusions: Modern pacemakers are well shielded against external EMFs, and workers with a pacemaker can most often return to their previous work after having a pacemaker implanted. However, an appropriate risk assessment is still necessary after the implantation of a pacemaker, a change of its generator, or major modification of its programming settings.

  1. Patients exposure from fluoroscopic guided pacemaker implantation procedures

    International Nuclear Information System (INIS)

    Alkhorayef, M.; Babikir, E.; Sulieman, A.; Daar, E.; Alnaaimi, M.; Alduaij, M.; Bradley, D.

    2016-10-01

    A pacemaker, which is used for heart re synchronization with electrical impulses, is used to manage many clinical conditions. Recently, the frequency of the pacemaker implantation procedures increased 50% worldwide. During this procedure, patients and staff can be exposed to excessive radiation exposure. Wide range of doses was reported in previous studies, suggesting that optimization of this procedure is not fulfilled yet. This study aims to evaluate the patient and staff radiation doses during cardiac pacemaker procedure and quantify the patient effective dose. A total of 145 procedures were performed for five pacemakers procedures (VVI, VVIR, VVD, VVDR and DDDR) two hospitals were evaluated. Patients doses were measured using the kerma-area product meter. Effective doses were estimated using software based on Monte Carlo simulation from National Radiological Protection Board. The effective dose values were used to estimate the cancer risk from pacemaker procedure. Patients demographic data, exposure parameters for both fluoroscopy and radiography were quantified. The mean patients doses (Gy. cm 2 ) for VVI, VVIR, VVD, VVDR and DDDR was 1.52±0.13 (1.43-1.61), 3.28±2.34 (0.29-8.73), 3.04±1.67 (1.57-4.86), 6.04±2.326 and 19.2±3.6 (5.43-30.2), respectively, per procedure. The overall patients effective dose is 1.1 mSv per procedure. (Author)

  2. Patients exposure from fluoroscopic guided pacemaker implantation procedures

    Energy Technology Data Exchange (ETDEWEB)

    Alkhorayef, M.; Babikir, E. [King Saud University, College of Applied Sciences, Radiological Sciences Department, P. O. Box 10219, Riyadh 11433 (Saudi Arabia); Sulieman, A. [Prince Sattam bin Abdulaziz University, College of Applied Medical Sciences, Radiology and Medical Imaging Department, P. O. Box 422, Alkharj 11942 (Saudi Arabia); Daar, E. [University of Jordan, Faculty of Science, Department of Physics, Amman 11942 (Jordan); Alnaaimi, M.; Alduaij, M. [Kuwait Cancer Control Centre, Department of Nuclear Medicine, Shwiekh (Kuwait); Bradley, D., E-mail: malkhorayef@ksu.edu.sa [University of Surrey, Centre for Nuclear and Radiation Physics, Guildford, Surrey, GU2 7XH (United Kingdom)

    2016-10-15

    A pacemaker, which is used for heart re synchronization with electrical impulses, is used to manage many clinical conditions. Recently, the frequency of the pacemaker implantation procedures increased 50% worldwide. During this procedure, patients and staff can be exposed to excessive radiation exposure. Wide range of doses was reported in previous studies, suggesting that optimization of this procedure is not fulfilled yet. This study aims to evaluate the patient and staff radiation doses during cardiac pacemaker procedure and quantify the patient effective dose. A total of 145 procedures were performed for five pacemakers procedures (VVI, VVIR, VVD, VVDR and DDDR) two hospitals were evaluated. Patients doses were measured using the kerma-area product meter. Effective doses were estimated using software based on Monte Carlo simulation from National Radiological Protection Board. The effective dose values were used to estimate the cancer risk from pacemaker procedure. Patients demographic data, exposure parameters for both fluoroscopy and radiography were quantified. The mean patients doses (Gy. cm{sup 2}) for VVI, VVIR, VVD, VVDR and DDDR was 1.52±0.13 (1.43-1.61), 3.28±2.34 (0.29-8.73), 3.04±1.67 (1.57-4.86), 6.04±2.326 and 19.2±3.6 (5.43-30.2), respectively, per procedure. The overall patients effective dose is 1.1 mSv per procedure. (Author)

  3. Rhabdomyosarcoma associated with the lead wire of a pacemaker generator implant.

    Science.gov (United States)

    Thieman Mankin, Kelley M; Dunbar, Mark D; Toplon, David; Ginn, Pamela; Maisenbacher, Herbert W; Risselada, Marije

    2014-06-01

    An 11-year-old female spayed Labrador Retriever was presented for a draining, painful subcutaneous mass palpated over a previously implanted pacemaker generator. Infection was suspected and the mass was removed surgically. On cut surface, the mass was friable and mottled tan to brown with firm pale tan nodules, surrounding the pacemaker lead wire adjacent to the pacemaker generator. Cytologic interpretation of impression smears was consistent with a sarcoma, and suggestive of a rhabdomyosarcoma due to the presence of strap-like cells. On histopathologic examination, a highly invasive nodular mass surrounded the pacemaker lead, composed of pleomorphic round, spindle and strap cells, and multinucleated giant cells. The population exhibited microscopic invasion into the deep portion of the fibrous capsule surrounding the pacemaker generator. There were tumor emboli within small to medium subcutaneous veins adjacent to the mass. Immunohistochemically, the neoplastic cells stained positive for α-sarcomeric actin and vimentin, and negative for α-smooth muscle actin, consistent with a rhabdomyosarcoma arising at the site of the pacemaker generator. To our knowledge, this is the first report of a rhabdomyosarcoma associated with the lead wire of a pacemaker generator in a dog. © 2014 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology.

  4. Effects of irradiation on the components of implantable pacemakers

    International Nuclear Information System (INIS)

    Kawamura, Shinji; Ono, Seiji; Kuga, Noriyuki; Shiba, Tooru; Hirose, Tetsuo; Matoba, Masaru

    2003-01-01

    The purpose of this study was to examine the effects of irradiation on implantable pacemaker components. The pacemaker was divided into three components: lead wire and electrode, battery, and electrical circuit, and each component was irradiated by X-ray and electron beams, respectively. The pacemaker parameters were measured by both telemetry data of the programmer and directly measured data from the output terminal. The following results were obtained. For the lead wire and electrode, there was no effect on the pacemaker function due to irradiation by X-ray and electron beams. In the case of battery irradiation, there was no change in battery voltage or current up to 236 Gy X-ray dose. In the electrical circuit, the pacemaker reverted to the regular beating rate (fixed-rate mode) immediately after the start of X-ray irradiation, and it continued in this mode during irradiation. In patients with their own heartbeat rhythm, changing to the fixed-rate mode may cause dangerous conditions such as ventricular fibrillation. When the accumulated irradiation dose is increased, another failure can be seen in the output voltage of the pacemaker. The pacing output voltage dropped rapidly by about 40% at 30-88 Gy. Decreasing the output voltage results in pacing disorders, and heart failure may occur. In the telemetry data of the programmer, no change in output voltage could be detected, highlighting the difference between telemetry data and actual pacing data. (author)

  5. The influence of electromagnetic interference and ionizing radiation on cardiac pacemakers

    International Nuclear Information System (INIS)

    Salmi, J.; Malmivuo, J.A.V.

    1990-01-01

    Adverse effects of the ionizing and non-ionizing electromagnetic fields on five pacemaker models have been tested. The study consisted of three parts: 1. measurement of magnetic fields in a radiotherapy room (microtron MM14), 2. the application of non-ionizing electromagnetic fields on pacemakers in a test laboratory (1 ... 1000 μT, 10 ... 10 000 Hz), and 3. the application of ionizing radiation of different types of radiotherapy devices on the pacemakers. The magnetic field strength in the microtron treatment room was found to be under 7.5 μT, which is one order of magnitude lower than the tolerance level obtained for the pacemakers in the test laboratory. All the tested pacemakers tolerated the ionizing radiation dose levels (less than 60 Gy) which are used in the radiotherapy. (orig.) [de

  6. Conduction Abnormalities and Pacemaker Implantations After SAPIEN 3 Vs SAPIEN XT Prosthesis Aortic Valve Implantation.

    Science.gov (United States)

    Husser, Oliver; Kessler, Thorsten; Burgdorf, Christof; Templin, Christian; Pellegrini, Costanza; Schneider, Simon; Kasel, Albert Markus; Kastrati, Adnan; Schunkert, Heribert; Hengstenberg, Christian

    2016-02-01

    Transcatheter aortic valve implantation is increasingly used in patients with aortic stenosis. Post-procedural intraventricular conduction abnormalities and permanent pacemaker implantations remain a serious concern. Recently, the Edwards SAPIEN 3 prosthesis has replaced the SAPIEN XT. We sought to determine the incidences of new-onset intraventricular conduction abnormalities and permanent pacemaker implantations by comparing the 2 devices. We analyzed the last consecutive 103 patients undergoing transcatheter aortic valve implantation with SAPIEN XT before SAPIEN 3 was used in the next 105 patients. To analyze permanent pacemaker implantations and new-onset intraventricular conduction abnormalities, patients with these conditions at baseline were excluded. Electrocardiograms were recorded at baseline, after the procedure, and before discharge. SAPIEN 3 was associated with higher device success (100% vs 92%; P=.005) and less paravalvular leakage (0% vs 7%; Ppacemaker implantations was 12.6% (23 of 183) with no difference between the 2 groups (SAPIEN 3: 12.5% [12 of 96] vs SAPIEN XT: 12.6% [11 of 87]; P=.99). SAPIEN 3 was associated with a higher rate of new-onset intraventricular conduction abnormalities (49% vs 27%; P=.007) due to a higher rate of fascicular blocks (17% vs 5%; P=.021). There was no statistically significant difference in transient (29% [20 of 69] vs persistent 19% [12 of 64]; P=.168) left bundle branch blocks (28% [19 of 69] vs 17% [11 of 64]; P=.154) when SAPIEN 3 was compared with SAPIEN XT. We found a trend toward a higher rate of new-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT, although this did not result in a higher permanent pacemaker implantation rate. Copyright © 2015 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.

  7. Dissecting aneurysm of arch and descending thoracic aorta presenting as a left sided hemorrhagic pleural effusion

    Directory of Open Access Journals (Sweden)

    Shamim Shelley

    2010-01-01

    Full Text Available The most common cause of massive hemorrhagic effusion is malignancy. Herein we present a case of dissecting aneurysm of descending thoracic aorta presenting initially with shortness of breath due to left sided massive pleural effusion. Effusion was hemorrhagic in nature with high hematocrit value. CT scan of thorax with CT angiogram was done and that revealed the diagnosis.

  8. Complications of pacemaker therapy in adults with congenital heart disease: a multicenter study.

    Science.gov (United States)

    Opić, Petra; van Kranenburg, Matthijs; Yap, Sing-Chien; van Dijk, Arie P; Budts, Werner; Vliegen, Hubert W; van Erven, Lieselot; Can, Anil; Sahin, Gulhan; Theuns, Dominic A M J; Witsenburg, Maarten; Roos-Hesselink, Jolien W

    2013-10-09

    This study aims to investigate indications and complications of permanent cardiac pacing in adults with congenital heart disease (CHD). Two-hundred and seventy-four CHD patients were identified who underwent permanent pacemaker implantation between 1972 and 2009. The indication for pacing was acquired sinus node or AV node conduction disease (63%), sinus node or AV node conduction disease after cardiac surgery (28%), and drug/arrhythmia-related indications (9%). Patients with complex CHD received a pacemaker at younger age (23 versus 31 years, ppacemaker implantation (general population: 5.2%). The most common acute complications were lead dysfunction (4.0%), bleeding (2.6%), pocket infection (1.5%) and pneumothorax (1.5%). During a median follow-up of 12 years, pacemaker-related complications requiring intervention occurred in 95 patients (34.6%). The most common late pacemaker-related complications included lead failure (24.8%), pacemaker dysfunction/early battery depletion (5.1%), pacemaker migration (4.7%) and erosion (4.7%). Pacemaker implantation at younger age (pacemaker-related complication (adjusted hazard ratio 1.68, 95% confidence interval 1.07 to 2.63, p=0.023). The risk of periprocedural complications seems higher in the CHD population compared to the general population and more than one-third of CHD patients encountered a pacemaker-related complication during long-term follow-up. This risk increases for those who receive a pacemaker at younger age. Crown Copyright © 2013. Published by Elsevier Ireland Ltd. All rights reserved.

  9. Interference of apex locator, pulp tester and diathermy on pacemaker function.

    Science.gov (United States)

    Sriman, Narayanan; Prabhakar, V; Bhuvaneswaran, J S; Subha, N

    2015-01-01

    The purpose of this study was to evaluate the effects of three electronic apex locators (EAL), electric pulp tester (EPT) and diathermy on pacemaker function in vitro. Three EALs: Root ZX (J. Morita Co., Tustin, CA, U.S.A.), Propex (Dentsply), Mini Apex locator (SybronEndo, Anaheim, CA, USA), EPT (Parkell pulp vitality tester Farmingdale, NY, USA) and Diathermy (Neomed 250 B) were tested for any interference with one pacemaker (A medtronic kappa KVDD901-serial number: PLE734632S). Directly connecting the pacemaker lead with the EAL/EPT/diathermy operating on a flat bench top, the telemetry wand was held directly over the pacemaker to monitor the pacing pattern for a period of 30 s. Pacemaker activity was continuously recorded on the telemetric programmer and electro gram (EGM) readings examined for pacer inhibition, noise reversion or inappropriate pacemaker pulses. All the three apex locators showed no pacing interference or background noise during its function or at rest. The EGM readings of EPT showed varying levels of background noise in between pacing however, this did not affect the normal pacing pattern and the pacing interval remained constant. EGM readings of diathermy showed an increase in the pacing interval (irregular pacing pattern) followed by complete inhibition of the pacing system. The tested EALs do not interfere with cardiac pacemaker function. The tested EPT showed varying levels of background noise but does not interfere with cardiac pacemaker function. Use of Diathermy interfered with the normal pacing, leading to complete inhibition of the pacing system.

  10. Deactivation of Pacemaker: Ethical Approach or Managerial Failure?

    Directory of Open Access Journals (Sweden)

    Macková Marie

    2017-12-01

    Full Text Available The decision about the deactivation of a pacemaker must be the result of a multicriteria decision-making process where the legal, ethical and effectiveness aspects must be taken into account and delicately balanced, while also considering the risk of managerial failure. Academic as well as professional discussion is necessary because there is a whole range of question marks on this topic and all the aspects mentioned above. The aim of this paper is to contribute to the debate by presenting the views of Czech physicians about the possibility of deactivation of the pacemaker in patients in terminal states. Based on the results of our research, the following steps are recommended to enable the deactivation of pacemakers in the Czech environment. Before the patient’s own indication of pacemaker therapy, treatment should be discussed with the patient in detail, including complications and deactivation options. Czech ethical consultant services should be set up in Czech hospitals. And last but not least, they should take an opinion on this issue as well as the professional society.

  11. Possible health hazards for cardiac pacemaker wearers from exposure to electromagnetic fields

    Energy Technology Data Exchange (ETDEWEB)

    1988-03-01

    Cardiac pacemakers are used to provide electrical stimulation to the heart when the heart's natural rhythm is interrupted. This study shows that they can be susceptible to electromagnetic fields. Pacemakers are well protected against common electromagnetic fields, such as those from household appliances. But intense electomagnetic fields, such as those found in some industrial settings, could affect the functioning of the pacemaker. Such interference may cause the pacemaker wearer to feel dizzy or experience an accelerated heartbeat. While this is not fatal, the pacemaker wearer should try to move away from the source of the interfering field and avoid situations in which interference could arise. After experiencing any of these symptoms, the pacemaker wearer should contact a physician. Potential sources of electromagnetic interference should be identified and characterized to determine if there could be an interference hazard. Exposure to interfering electomagnetic fields should be minimized. 7 refs., 1 fig.

  12. Swim pacemakers in box jellyfish are modulated by the visual input

    DEFF Research Database (Denmark)

    Garm, Anders Lydik; Bielecki, Jan

    2008-01-01

    A major part of the cubozoan central nervous system is situated in the eye-bearing rhopalia. One of the neuronal output channels from the rhopalia carries a swim pacemaker signal, which has a one-to-one relation with the swim contractions of the bell shaped body. Given the advanced visual system...... of box jellyfish and that the pacemaker signal originates in the vicinity of these eyes, it seems logical to assume that the pacemakers are modified by the visual input. Here, the firing frequency and distribution of inter-signal intervals (ISIs) of single pacemakers are examined in the Caribbean box...

  13. State of the art of cardiac pacemaker technology

    International Nuclear Information System (INIS)

    Lambeck, R.

    1978-01-01

    The development of cardiac pacemakers from fixed-frequency to demand pacemakers is reviewed. The latter is described in more detail with regard to its energy sources and its design. The use of radioactive energy sources is illustrated by the example of 238 Pu and 147 promethium and a comparison of the two radiation sources. (AJ) 891 AJ [de

  14. Gender specific pattern of left ventricular cardiac adaptation to ...

    African Journals Online (AJOL)

    EB

    2013-09-03

    Sep 3, 2013 ... Some gender specific associations with left ventricular structure and function have been described ... used for analysis. ... risk due to LVH and that cardiac adaptation to ... had history taking, physical examination and ..... between the gender specific differences in cardiac ... A metaanalysis of individual.

  15. Cardiac pacemaker dysfunction in children after thoracic drainage catheter manipulation.

    Science.gov (United States)

    Lobdell, K W; Walters, H L; Hudson, C; Hakimi, M

    1997-05-01

    Two children underwent placement of permanent, epicardial-lead, dual-chamber, unipolar pacemaker systems for complete heart block. Postoperatively, both patients demonstrated subcutaneous emphysema-in the area of their pulse generators-temporally related to thoracic catheter manipulation. Acutely, each situation was managed with manual compression of the pulse generator, ascertaining appropriate pacemaker sensing and pacing. Maintenance of compression with pressure dressings, vigilant observation/monitoring, and education of the care givers resulted in satisfactory pacemaker function without invasive intervention.

  16. Acute pericarditis with cardiac tamponade induced by pacemaker implantation.

    Science.gov (United States)

    Shingaki, Masami; Kobayashi, Yutaka; Suzuki, Haruo

    2015-11-01

    An 87-year-old woman was diagnosed with third-degree atrioventricular block and underwent pacemaker implantation. On postoperative day 12, she experienced cardiac tamponade that was suspected on computed tomography to be caused by lead perforation; therefore, we performed open-heart surgery. However, we could not identify a perforation site on the heart, and drained a 400-mL exudative pericardial effusion. Subsequently, we diagnosed the pericardial effusion as due to pericarditis induced by pacemaker implantation. It is sometimes difficult to distinguish pericarditis from pacemaker lead perforation, so both should be included in the differential diagnosis. © The Author(s) 2014.

  17. Is pacemaker therapy the right key to patients with vasovagal syncope?

    Science.gov (United States)

    Radovanović, Nikola N; Kirćanski, Bratislav; Raspopović, Srdjan; Pavlović, Siniša U; Jovanović, Velibor; Milašinović, Goran

    2016-01-01

    Vasovagal syncope is the most common type of reflex syncope. Efficacy of cardiac pacing in this indication has not been the subject of many studies and pacemaker therapy in patients with vasovagal syncope is still controversial. This study aimed to assess the efficacy and safety of pacing therapy in treatment of patients with vasovagal syncope, to determine contribution of new therapeutic models in increasing its success, and to identify risk factors associated with a higher rate of symptoms after pacemaker implantation. A retrospective study included 30 patients with pacemaker implanted due to vasovagal syncope in the Pacemaker Center, Clinical Center of Serbia, between November 2003 and June 2014. Head-up tilt test was performed to diagnose vasovagal syncope. Patients with cardioinhibitory and mixed type of disease were enrolled in the study. Mean age was 48.1 ± 11.1 years and 18 (60%) patients were men. Mean follow-up period was 5.9 ± 3.0 years. Primarily, implantable loop recorder was implanted in 10 (33.3%) patients. Twenty (66.7%) patients presented cardioinhibitory and 10 (33.3%) mixed type of vasovagal syncope. After pacemaker implantation, 11 (36.7%) patients had syncope. In multiple logistic regression analysis we showed that syncope is statistically more likely to occur after pacemaker implantation in patients with mixed type of vasovagal syncope (p = 0.018). There were two (6.7%) perioperative surgical complications. Pacemaker therapy is a safe treatment for patients with vasovagal syncope, whose efficacy can be improved by strict selection of patients. We showed that symptoms occur statistically more often in patients with mixed type of disease after pacemaker implantation.

  18. Current use of implantable electrical devices in Sweden: data from the Swedish pacemaker and implantable cardioverter-defibrillator registry.

    Science.gov (United States)

    Gadler, Fredrik; Valzania, Cinzia; Linde, Cecilia

    2015-01-01

    The National Swedish Pacemaker and Implantable Cardioverter-Defibrillator (ICD) Registry collects prospective data on all pacemaker and ICD implants in Sweden. We aimed to report the 2012 findings of the Registry concerning electrical devices implantation rates and changes over time, 1 year complications, long-term device longevity and patient survival. Forty-four Swedish implanting centres continuously contribute implantation of pacemakers and ICDs to the Registry by direct data entry on a specific website. Clinical and technical information on 2012 first implants and postoperative complications were analysed and compared with previous years. Patient survival data were obtained from the Swedish population register database. In 2012, the mean pacemaker and ICD first implantation rates were 697 and 136 per million inhabitants, respectively. The number of cardiac resynchronization therapy (CRT) first implantations/million capita was 41 (CRT pacemakers) and 55 (CRT defibrillators), with only a slight increase in CRT-ICD rate compared with 2011. Most device implantations were performed in men. Complication rates for pacemaker and ICD procedures were 5.3 and 10.1% at 1 year, respectively. Device and lead longevity differed among manufacturers. Pacemaker patients were older at the time of first implant and had generally worse survival rate than ICD patients (63 vs. 82% after 5 years). Pacemaker and ICD implantation rates seem to have reached a level phase in Sweden. Implantable cardioverter-defibrillator and CRT implantation rates are very low and do not reflect guideline indications. Gender differences in CRT and ICD implantations are pronounced. Device and patient survival rates are variable, and should be considered when deciding device type. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

  19. Multiple photoreceptor systems control the swim pacemaker activity in box jellyfish

    DEFF Research Database (Denmark)

    Garm, Anders Lydik; Mori, S.

    2009-01-01

    Like all other cnidarian medusae, box jellyfish propel themselves through the water by contracting their bell-shaped body in discrete swim pulses. These pulses are controlled by a swim pacemaker system situated in their sensory structures, the rhopalia. Each medusa has four rhopalia each with a s......Like all other cnidarian medusae, box jellyfish propel themselves through the water by contracting their bell-shaped body in discrete swim pulses. These pulses are controlled by a swim pacemaker system situated in their sensory structures, the rhopalia. Each medusa has four rhopalia each...... with a similar set of six eyes of four morphologically different types. We have examined how each of the four eye types influences the swim pacemaker. Multiple photoreceptor systems, three of the four eye types, plus the rhopalial neuropil, affect the swim pacemaker. The lower lens eye inhibits the pacemaker...... when stimulated and provokes a strong increase in the pacemaker frequency upon light-off. The upper lens eye, the pit eyes and the rhopalial neuropil all have close to the opposite effect. When these responses are compared with all-eye stimulations it is seen that some advanced integration must take...

  20. Continued evaluation of cardiac pacemakers. Annual report Jun 73-31 Aug 75

    International Nuclear Information System (INIS)

    Brueschke, E.E.; Uretz, E.F.; Hauser, R.G.

    1976-01-01

    The nuclear powered pacemaker evaluation was designed to characterize the operating characteristics of such pacemakers and identify potential failure mechanisms. More than 30 nuclear powered pacemakers were implanted in dogs and 15 were subjected to high stress bench tests. This evaluation resulted in the identification of several basic pacemaker problem areas for the batteries under test. These included abrupt and gradual nuclear battery failures, weld seam defects (in one model) resulting in pulse generator failures, pulse generator lead connector defects and pulse generator electronic malfunctions

  1. I-123 metaiodobenzylguanidine cardiac scintigraphy in patients with an implanted permanent pacemaker

    International Nuclear Information System (INIS)

    Nakata, Akio; Hirota, Satoshi; Tsuji, Hiroshi; Takazakura, Eisuke

    1995-01-01

    Tl scintigraphic abnormalities have been reported in patients with an implanted permanent pacemaker, but little is known about the MIBG scintigraphic findings in such patients. This study was performed to assess the MIBG scintigraphic findings in patients with an implanted permanent pacemaker, and to test the hypothesis that imaging characteristics of MIBG scintigraphy differ according to its mode. Twelve patients (4 men and 8 women, mean age: 72.4±9.5 years), who had undergone the implantation of a permanent pacemaker for bradyarrhythmias, underwent MIBG scintigraphy. The patients were divided into VVI pacemaker and DDD pacemaker groups. The tomograms were divided into nine segments and the MIBG defect in each segment scored on a scale ranging from 0 (normal uptake) to 3 (no uptake). Total MIBG defect scores were generated by summing the scores for the nine segments in each patient. MIBG scintigraphic abnormalities were found in ten of the twelve patients. The six patients with the VVI pacemaker manifested MIBG scintigraphic abnormalities. These MIBG scintigraphic abnormalities were observed in all segments, particularly in the posterior segments. The mean total defect score of the VVI group was higher than that of the DDD group (14.8±9.8 vs 3.0±3.5, respectively p<0.05). Therefore, we conclude that despite several limitations of the study, MIBG scintigraphic abnormalities occur in patients with implanted permanent pacemakers, and that such abnormalities are more prominent with the VVI than DDD pacemaker. (author)

  2. Imaging analysis of heart movement for improving the respiration-gated radiotherapy in patients with left sided breast cancer

    Energy Technology Data Exchange (ETDEWEB)

    Abdelhamid, Rania; Farrag, A.; Khalifa, A. [Clinical Oncology Department, Assiut University (Egypt); Block, Andreas [Institut fuer Medizinische Strahlenphysik und Strahlenschutz, Klinikum Dortmund (Germany)

    2012-07-01

    Respiration induced heart movement during radiotherapy exposes the heart to the inevitable risks of radio-exposure, and hence radiation injury, in cases of Lt. sided breast cancer. The impact of such a risk is additionally aggravated by the use of radiotherapy in combination with cardiotoxic chemotherapeutic agents. Radio-oncologists pay special attention to the coronary arteries that might be included in this small part of the heart exposed to radiation. The aim of this study was to include the internal heart movement for improving respiration-gated radiotherapy of left sided breast cancer. For 70 patients, all females left sided breast cancer, two planning CT's in inspiration and expiration, and one free breathing scan are performed. The heart motion was analyzed with the clinic-developed software ORAT in the simulator sequence for acquiring information of the cranio-caudal amplitude of heart movements in free breathing (respiration-induced amplitude) and a 15 seconds breath-hold phase (inherent amplitude). The role of inherent heart movement varies from one patient to another which should be taken in consideration during defining the parameters of respiration-gated radiotherapy. The inherent amplitude of the heart motion is the physiological lower limit of the respiration-gating window.

  3. Subjective consequences of permanent pacemaker therapy in patients under the age of retirement

    DEFF Research Database (Denmark)

    Mickley, H; Petersen, J; Nielsen, B L

    1989-01-01

    During a 5-year period, 81 patients ages 20 to 60 years old had implantation of a permanent cardiac pacemaker at the University Hospital, Odense. At follow-up, during 1985, the 73 survivors received a semi-structured questionnaire regarding subjective consequences of pacemaker therapy, and 72...... people (98.6%) agreed to participate. The mean pacing period (range) was 33.8 (11-72) months. Surgical intervention was required in 14 patients (19.4%) during follow-up. Regarding all symptoms 67 patients (93.1%) perceived benefit from the pacemaker. The effectiveness of cardiac pacing was most...... or a sensation of "impulses"/palpitations. To the majority (49 patients or 68.1%) pacemaker treatment did not influence quality of sexual activity. Six patients (8.3%) perceived an improvement, whereas a corresponding number felt deterioration in sexual activity following pacemaker implantation. Pacemaker...

  4. Long-term outcomes after resection of para-aortic lymph node metastasis from left-sided colon and rectal cancer.

    Science.gov (United States)

    Nakai, Nozomu; Yamaguchi, Tomohiro; Kinugasa, Yusuke; Shiomi, Akio; Kagawa, Hiroyasu; Yamakawa, Yushi; Numata, Masakatsu; Furutani, Akinobu

    2017-07-01

    Para-aortic lymph node (PALN) metastasis from colorectal cancer is rare and often not suitable for surgery. However, in selected patients, radical resection may bring about longer survival. The aim of this study was to evaluate long-term outcomes of resection of left-sided colon or rectal cancer with simultaneous PALN metastasis. The study included 2122 patients with left-sided colon or rectal cancer (30 patients with and 2092 patients without PALN metastasis) who underwent resection with curative intent between 2002 and 2013. Clinicopathological characteristics, long-term outcomes of resection, and factors related to poor postoperative survival in patients with PALN metastasis were investigated. Of a total of 2122 total patients, 16 of 50 patients (32.0%) with lymph node metastasis at the root of the inferior mesenteric artery had PALN metastasis. The 5-year overall survival rates for 18 patients who underwent R0 resection and 12 patients who did not were 29.1 and 10.4%, respectively (p = 0.017). Factors associated with poor postoperative survival among patients who underwent R0 resection were presence of conversion therapy, lack of adjuvant chemotherapy, carcinoembryonic antigen >20 ng/mL, and lateral lymph node metastasis in rectal cancer patients. The 5-year recurrence-free survival rate was 14.8%. Although recurrence was frequent, R0 resection for left-sided colon or rectal cancer with PALN metastasis was associated with longer survival than R1/R2 resection. Furthermore, the 5-year overall survival rate in the R0 group was relatively favorable for stage IV. Therefore, R0 resection may prolong survival compared with chemotherapy alone in selected patients.

  5. Zero-fluoroscopy permanent pacemaker implantation using Ensite NavX system: Clinical viability or fanciful technique?

    Science.gov (United States)

    Guo, Ping; Qiu, Jie; Wang, Yan; Chen, Guangzhi; Proietti, Riccardo; Fadhle, Al-Selmi; Zhao, Chunxia; Wen Wang, Dao

    2018-02-01

    Fluoroscopy is the imaging modality routinely used for cardiac device implantation and electrophysiological procedures. Due to the rising concern regarding the harmful effects of radiation exposure to both the patients and operation staffs, novel 3D mapping systems have been developed and implemented in electrophysiological procedure for the navigation of catheters inside the heart chambers. Their applicability in cardiac device implantation has been rarely reported. Our aim is to evaluate the feasibility and safety of permanent pacemaker implantation without fluoroscopy. From January 2012 to June 2016, six patients (50 ± 15 years, four of six were female, one of who was at the 25th week of gestation) who underwent permanent pacemaker implantation were included (zero-fluoroscopy group). Data from 20 consecutive cases of implantation performed under fluoroscopy guidance were chosen as a control group (fluoroscopy group). Total implantation procedure time for single-chamber pacemaker was 51.3 ± 13.1 minutes in the zero-fluoroscopy group and 42.6 ± 7.4 minutes in the fluoroscopy group (P  =  0.155). The implantation procedural time for a dual-chamber pacemaker was 88.3 ± 19.6 minutes and 67.3 ± 7.6 minutes in the zero-fluoroscopy and fluoroscopy groups (P  =  0.013), respectively. No complications were observed during the procedure and the follow-up in the two groups, and all pacemakers worked with satisfactory parameters. Ensite NavX system can be used as a reliable and safe zero-fluoroscopy approach for the implantation of single- or dual-chamber permanent pacemakers in specific patients, such as pregnant women or in extreme situations when the x-ray machine is not available. © 2017 The Authors. Pacing and Clinical Electrophysiology published by Wiley Periodicals, Inc.

  6. An unusual case of metastasis to the left side of the heart: a case report

    Directory of Open Access Journals (Sweden)

    Boyars Michael

    2011-01-01

    Full Text Available Abstract Introduction Cardiac metastases are found in six to 20% of autopsies of patients with malignant neoplasm. The most common neoplasms that metastasize to the heart are malignant melanoma, lymphoma, and leukemia, but the relative numbers are greater with breast and lung cancers, reflecting the most common incidence of these cancers. Case presentation A 60-year-old Hispanic man presented to our hospital after being transferred from an outside hospital for workup and evaluation of an adrenal mass of the abdomen and pelvis, found on computed tomography. His chief complaint upon admission was altered mental status. Physical examination was unremarkable. He was alert and oriented and had a dry and non-erythematous oropharynx, and bilateral diffuse wheezing on lung examination. Computed tomography of the chest showed multiple hypodense lesions in the left ventricular myocardium, suggestive of metastases. There were also tiny sub-centimeter nodular densities in the right upper and lower lobes. Adrenal glands contained hypodense lesions, which showed characteristic adenocarcinomatous malignant cells. Conclusion Cancers which have metastasized to the heart are found in six to 20% of patients with malignant neoplasms. The right side of the heart is more commonly involved in metastasis. This study is unusual in that a tumor of an unknown primary origin had metastasized to the left side of the heart.

  7. Twiddler-syndrom er en årsag til pacemaker elektrode displacering

    DEFF Research Database (Denmark)

    Grønbech, Keea Treu; Hansen, Michael Gilså

    2013-01-01

    Twiddler's syndrome is a rare cause of pacemaker electrode displacement. The displacement is caused by the patient's manipulation with the pacemaker, so the electrode is retracted. We describe a case of a 79-year-old overweight woman with a known psychiatric anamnesis, who was admitted twice...... with twiddler's syndrome. Age and overweight are known risk factors for twiddler's syndrome; and in this case the patient's psychiatric habitus was probably an additional risk factor. Before performing a pacemaker implantation it is important to take the patient's risk factors into account, and thus consider...

  8. Respiration Induced Heart Motion and Indications of Gated Delivery for Left-Sided Breast Irradiation

    International Nuclear Information System (INIS)

    Qi, X. Sharon; Hu, Angela; Wang Kai; Newman, Francis; Crosby, Marcus; Hu Bin; White, Julia; Li, X. Allen

    2012-01-01

    Purpose: To investigate respiration-induced heart motion for left-sided breast irradiation using a four-dimensional computed tomography (4DCT) technique and to determine novel indications to assess heart motion and identify breast patients who may benefit from a gated treatment. Methods and Materials: Images of 4DCT acquired during free breathing for 20 left-sided breast cancer patients, who underwent whole breast irradiation with or without regional nodal irradiation, were analyzed retrospectively. Dose distributions were reconstructed in the phases of 0%, 20%, and 50%. The intrafractional heart displacement was measured in three selected transverse CT slices using D LAD (the distance from left ascending aorta to a fixed line [connecting middle point of sternum and the body] drawn on each slice) and maximum heart depth (MHD, the distance of the forefront of the heart to the line). Linear regression analysis was used to correlate these indices with mean heart dose and heart dose volume at different breathing phases. Results: Respiration-induced heart displacement resulted in observable variations in dose delivered to the heart. During a normal free-breathing cycle, heart-induced motion D LAD and MHD changed up to 9 and 11 mm respectively, resulting in up to 38% and 39% increases of mean doses and V 25.2 for the heart. MHD and D LAD were positively correlated with mean heart dose and heart dose volume. Respiratory-adapted gated treatment may better spare heart and ipsilateral-lung compared with the conventional non-gated plan in a subset of patients with large D LAD or MHD variations. Conclusion: Proposed indices offer novel assessment of heart displacement based on 4DCT images. MHD and D LAD can be used independently or jointly as selection criteria for respiratory gating procedure before treatment planning. Patients with great intrafractional MHD variations or tumor(s) close to the diaphragm may particularly benefit from the gated treatment.

  9. Respiration Induced Heart Motion and Indications of Gated Delivery for Left-Sided Breast Irradiation

    Energy Technology Data Exchange (ETDEWEB)

    Qi, X. Sharon, E-mail: xiangrong.qi@ucdenver.edu [Department of Radiation Oncology, University of Colorado Denver, Aurora, CO (United States); Hu, Angela [Department of Radiation Oncology, University of Colorado Denver, Aurora, CO (United States); Wang Kai [Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI (United States); Newman, Francis [Department of Radiation Oncology, University of Colorado Denver, Aurora, CO (United States); Crosby, Marcus; Hu Bin; White, Julia; Li, X. Allen [Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI (United States)

    2012-04-01

    Purpose: To investigate respiration-induced heart motion for left-sided breast irradiation using a four-dimensional computed tomography (4DCT) technique and to determine novel indications to assess heart motion and identify breast patients who may benefit from a gated treatment. Methods and Materials: Images of 4DCT acquired during free breathing for 20 left-sided breast cancer patients, who underwent whole breast irradiation with or without regional nodal irradiation, were analyzed retrospectively. Dose distributions were reconstructed in the phases of 0%, 20%, and 50%. The intrafractional heart displacement was measured in three selected transverse CT slices using D{sub LAD} (the distance from left ascending aorta to a fixed line [connecting middle point of sternum and the body] drawn on each slice) and maximum heart depth (MHD, the distance of the forefront of the heart to the line). Linear regression analysis was used to correlate these indices with mean heart dose and heart dose volume at different breathing phases. Results: Respiration-induced heart displacement resulted in observable variations in dose delivered to the heart. During a normal free-breathing cycle, heart-induced motion D{sub LAD} and MHD changed up to 9 and 11 mm respectively, resulting in up to 38% and 39% increases of mean doses and V{sub 25.2} for the heart. MHD and D{sub LAD} were positively correlated with mean heart dose and heart dose volume. Respiratory-adapted gated treatment may better spare heart and ipsilateral-lung compared with the conventional non-gated plan in a subset of patients with large D{sub LAD} or MHD variations. Conclusion: Proposed indices offer novel assessment of heart displacement based on 4DCT images. MHD and D{sub LAD} can be used independently or jointly as selection criteria for respiratory gating procedure before treatment planning. Patients with great intrafractional MHD variations or tumor(s) close to the diaphragm may particularly benefit from the gated

  10. Myocardial perfusion changes in patients irradiated for left-sided breast cancer and correlation with coronary artery distribution

    International Nuclear Information System (INIS)

    Lind, Pehr A.; Pagnanelli, Robert; Marks, Lawrence B.; Borges-Neto, Salvador; Hu, Caroline; Zhou, S.-M.; Light, Kim; Hardenbergh, Patricia H.

    2003-01-01

    Purpose: To evaluate postradiation regional heart perfusion changes with single photon emission tomography (SPECT) myocardial perfusion imaging in 69 patients treated with tangential photon beams radiation therapy (RT) for left-sided breast cancer. To correlate SPECT changes with percent irradiated left ventricle (LV) volume and risk factors for coronary artery disease (CAD). Methods and Materials: Rest SPECT of the LV was acquired pre-RT and at 6-month intervals post-RT. The extent of defects (%) with a severity > 1.5 standard deviations below the mean was quantitatively analyzed for the distributions of the left anterior descending (LAD) artery, left circumflex (LCX) artery, and right coronary artery (RCA) based on computer assisted polar map reconstruction (i.e., bull's-eye-view). Changes in perfusion were correlated with percent irradiated LV receiving > 25 Gy (range 0-32%). Data on patient- and treatment-related factors were collected prospectively (e.g., cardiac premorbidity, risk factors for CAD, chemotherapy, and hormonal treatment). Results: In the LAD distribution, there were increased perfusion defects at 6 months (median 11%; interquartile range 2-23) compared with baseline (median 5%; interquartile range 1-14) (p<0.001). There were no increases in perfusion defects in the LCX or RCA distributions. In multivariate analysis, the SPECT perfusion changes in the LAD distribution at 6 months were independently associated with percent irradiated LV (p<0.001), hormonal therapy (p=0.005), and pre-RT hypercholesterolemia (p=0.006). The SPECT defects in the LAD distribution at 12 and 18 months were not statistically different from those at 6 months. The perfusion defects in the LAD distribution were limited essentially to the regions of irradiated myocardium. Conclusion: Tangential photon beam RT in patients with left-sided breast cancer was associated with short-term SPECT defects in the vascular distribution corresponding to the radiation portals. Factors

  11. Residue-Specific Side-Chain Polymorphisms via Particle Belief Propagation.

    Science.gov (United States)

    Ghoraie, Laleh Soltan; Burkowski, Forbes; Li, Shuai Cheng; Zhu, Mu

    2014-01-01

    Protein side chains populate diverse conformational ensembles in crystals. Despite much evidence that there is widespread conformational polymorphism in protein side chains, most of the X-ray crystallography data are modeled by single conformations in the Protein Data Bank. The ability to extract or to predict these conformational polymorphisms is of crucial importance, as it facilitates deeper understanding of protein dynamics and functionality. In this paper, we describe a computational strategy capable of predicting side-chain polymorphisms. Our approach extends a particular class of algorithms for side-chain prediction by modeling the side-chain dihedral angles more appropriately as continuous rather than discrete variables. Employing a new inferential technique known as particle belief propagation, we predict residue-specific distributions that encode information about side-chain polymorphisms. Our predicted polymorphisms are in relatively close agreement with results from a state-of-the-art approach based on X-ray crystallography data, which characterizes the conformational polymorphisms of side chains using electron density information, and has successfully discovered previously unmodeled conformations.

  12. Dosimetric perturbations due to an implanted cardiac pacemaker in MammoSite{sup Registered-Sign} treatment

    Energy Technology Data Exchange (ETDEWEB)

    Sung, Wonmo; Kim, Siyong; Kim, Jung-in; Lee, Jae-gi; Shin, Young-Joo; Jung, Jae-Yong; Ye, Sung-Joon [Interdisciplinary Program in Radiation Applied Life Science, Seoul National University College of Medicine, Seoul 110-799, South Korea and Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 110-744 (Korea, Republic of); Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida 32224 (United States); Interdisciplinary Program in Radiation Applied Life Science, Seoul National University College of Medicine, Seoul 110-799, South Korea and Department of Radiation Oncology, Kangbuk Samsung Medical Center, Seoul 110-746 (Korea, Republic of); Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 110-744, South Korea and Department of Biomedical Radiation Sciences, Seoul National University, Seoul 151-742 (Korea, Republic of); Department of Radiation Oncology, Sanggye Paik Hospital, Inje University, Seoul 139-707 (Korea, Republic of); Interdisciplinary Program in Radiation Applied Life Science, Seoul National University College of Medicine, Seoul 110-799 (Korea, Republic of); Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 110-744 (Korea, Republic of); Department of Biomedical Radiation Sciences, Seoul National University, Seoul 151-742 (Korea, Republic of); Department of Radiation Oncology, Seoul National University, Seoul 110-799 (Korea, Republic of) and Advanced Institutes of Convergence Technology, Seoul National University, Suwon 443-270 (Korea, Republic of)

    2012-10-15

    Purpose: To investigate dose perturbations for pacemaker-implanted patients in partial breast irradiation using high dose rate (HDR) balloon brachytherapy. Methods: Monte Carlo (MC) simulations were performed to calculate dose distributions involving a pacemaker in Ir-192 HDR balloon brachytherapy. Dose perturbations by varying balloon-to-pacemaker distances (BPD = 50 or 100 mm) and concentrations of iodine contrast medium (2.5%, 5.0%, 7.5%, and 10.0% by volume) in the balloon were investigated for separate parts of the pacemaker (i.e., battery and substrate). Relative measurements using an ion-chamber were also performed to confirm MC results. Results: The MC and measured results in homogeneous media without a pacemaker agreed with published data within 2% from the balloon surface to 100 mm BPD. Further their dose distributions with a pacemaker were in a comparable agreement. The MC results showed that doses over the battery were increased by a factor of 3, compared to doses without a pacemaker. However, there was no significant dose perturbation in the middle of substrate but up to 70% dose increase in the substrate interface with the titanium capsule. The attenuation by iodine contrast medium lessened doses delivered to the pacemaker by up to 9%. Conclusions: Due to inhomogeneity of pacemaker and contrast medium as well as low-energy photons in Ir-192 HDR balloon brachytherapy, the actual dose received in a pacemaker is different from the homogeneous medium-based dose and the external beam-based dose. Therefore, the dose perturbations should be considered for pacemaker-implanted patients when evaluating a safe clinical distance between the balloon and pacemaker.

  13. Intraoperative colonic pulse oximetry in left-sided colorectal surgery: can it predict anastomotic leak?

    Science.gov (United States)

    Salusjärvi, Johannes M; Carpelan-Holmström, Monika A; Louhimo, Johanna M; Kruuna, Olli; Scheinin, Tom M

    2018-03-01

    An anastomotic leak is a fairly common and a potentially lethal complication in colorectal surgery. Objective methods to assess the viability and blood circulation of the anastomosis could help in preventing leaks. Intraoperative pulse oximetry is a cheap, easy to use, fast, and readily available method to assess tissue viability. Our aim was to study whether intraoperative pulse oximetry can predict the development of an anastomotic leak. The study was a prospective single-arm study conducted between the years 2005 and 2011 in Helsinki University Hospital. Patient material consisted of 422 patients undergoing elective left-sided colorectal surgery. The patients were operated by one of the three surgeons. All of the operations were partial or total resections of the left side of the colon with a colorectal anastomosis. The intraoperative colonic oxygen saturation was measured with pulse oximetry from the colonic wall, and the values were analyzed with respect to post-operative complications. 2.3 times more operated anastomotic leaks occurred when the colonic StO 2 was ≤ 90% (11/129 vs 11/293). The mean colonic StO 2 was 91.1 in patients who developed an operated anastomotic leak and 93.0 in patients who did not. With logistic regression analysis, the risk of operated anastomotic leak was 4.2 times higher with StO 2 values ≤ 90%. Low intraoperative colonic StO 2 values are associated with the occurrence of anastomotic leak. Despite its handicaps, the method seems to be useful in assessing anastomotic viability.

  14. Semiconductor measurement technology: reliability technology for cardiac pacemakers 2: a workshop report, 1976

    International Nuclear Information System (INIS)

    Schafft, H.A.

    1977-01-01

    Summaries are presented of 12 invited talks on the following topics: the procurement and assurance of high reliability electronic parts, leak rate and moisture measurements, pacemaker batteries, and pacemaker leads. The workshop, second in a series, was held in response to strong interest expressed by the pacemaker community to address technical questions relevant to the enhancement and assurance of cardiac pacemaker reliability. Discussed at the workshop were a process validation wafer concept for assuring process uniformity in device chips; screen tests for assuring reliable electronic parts; reliability prediction; reliability comparison of semiconductor technologies; mechanisms of short-circuiting dendritic growths; details of helium and radioisotope leak test methods; a study to correlate package leak rates, as measured with test gasses, and actual moisture infusion; battery life prediction; microcalorimetric measurements to nondestructively evaluate batteries for pacemakers; and an engineer's and a physician's view of the present status of pacemaker leads. References are included with most of the reports

  15. Helical Tomotherapy Planning for Left-Sided Breast Cancer Patients With Positive Lymph Nodes: Comparison to Conventional Multiport Breast Technique

    International Nuclear Information System (INIS)

    Goddu, S. Murty; Chaudhari, Summer; Mamalui-Hunter, Maria; Pechenaya, Olga L.; Pratt, David; Mutic, Sasa; Zoberi, Imran; Jeswani, Sam; Powell, Simon N.; Low, Daniel A.

    2009-01-01

    Purpose: To evaluate the feasibility of using helical tomotherapy for locally advanced left-sided breast cancer. Methods and Materials: Treatment plans were generated for 10 left-sided breast cancer patients with positive lymph nodes comparing a multiport breast (three-dimensional) technique with the tomotherapy treatment planning system. The planning target volumes, including the chest wall/breast, supraclavicular, axillary, and internal mammary lymph nodes, were contoured. The treatment plans were generated on the tomotherapy treatment planning system to deliver 50.4 Gy to the planning target volume. To spare the contralateral tissues, directional blocking was applied to the right breast and right lung. The optimization goals were to protect the lungs, heart, and right breast. Results: The tomotherapy plans increased the minimal dose to the planning target volume (minimal dose received by 99% of target volume = 46.2 ± 1.3 Gy vs. 27.9 ± 17.1 Gy) while improving the dose homogeneity (dose difference between the minimal dose received by 5% and 95% of the planning target volume = 7.5 ± 1.8 Gy vs. 37.5 ± 26.9 Gy). The mean percentage of the left lung volume receiving ≥20 Gy in the tomotherapy plans decreased from 32.6% ± 4.1% to 17.6% ± 3.5%, while restricting the right-lung mean dose to <5 Gy. However, the mean percentage of volume receiving ≥5 Gy for the total lung increased from 25.2% ± 4.2% for the three-dimensional technique to 46.9% ± 8.4% for the tomotherapy plan. The mean volume receiving ≥35 Gy for the heart decreased from 5.6% ± 4.8% to 2.2% ± 1.5% in the tomotherapy plans. However, the mean heart dose for tomotherapy delivery increased from 7.5 ± 3.4 Gy to 12.2 ± 1.8 Gy. Conclusion: The tomotherapy plans provided better dose conformity and homogeneity than did the three-dimensional plans for treatment of left-sided breast tumors with regional lymph node involvement, while allowing greater sparing of the heart and left lung from doses

  16. Inverted Nutcracker Syndrome: A Case of Persistent Hematuria and Pain in the Presence of a Left-Sided Inferior Vena Cava

    Directory of Open Access Journals (Sweden)

    Obi Ekwenna

    2011-01-01

    Full Text Available Nutcracker syndrome is described as the symptomatic compression of left renal vein between the aorta and the superior mesenteric artery, resulting in outflow congestion of the left kidney. We present the case of a 51-year-old male with a left-sided inferior vena cava, resulting in compression of the right renal vein by the superior mesenteric artery. Secondary to this anatomic anomaly, the patient experienced a many-year history of flank pain and intermittent gross hematuria. We have termed this unusual anatomic finding and its associated symptoms as the “inverted nutcracker syndrome”, and describe its successful management with nephrectomy and autotransplantation.

  17. A fully implantable pacemaker for the mouse: from battery to wireless power.

    Science.gov (United States)

    Laughner, Jacob I; Marrus, Scott B; Zellmer, Erik R; Weinheimer, Carla J; MacEwan, Matthew R; Cui, Sophia X; Nerbonne, Jeanne M; Efimov, Igor R

    2013-01-01

    Animal models have become a popular platform for the investigation of the molecular and systemic mechanisms of pathological cardiovascular physiology. Chronic pacing studies with implantable pacemakers in large animals have led to useful models of heart failure and atrial fibrillation. Unfortunately, molecular and genetic studies in these large animal models are often prohibitively expensive or not available. Conversely, the mouse is an excellent species for studying molecular mechanisms of cardiovascular disease through genetic engineering. However, the large size of available pacemakers does not lend itself to chronic pacing in mice. Here, we present the design for a novel, fully implantable wireless-powered pacemaker for mice capable of long-term (>30 days) pacing. This design is compared to a traditional battery-powered pacemaker to demonstrate critical advantages achieved through wireless inductive power transfer and control. Battery-powered and wireless-powered pacemakers were fabricated from standard electronic components in our laboratory. Mice (n = 24) were implanted with endocardial, battery-powered devices (n = 14) and epicardial, wireless-powered devices (n = 10). Wireless-powered devices were associated with reduced implant mortality and more reliable device function compared to battery-powered devices. Eight of 14 (57.1%) mice implanted with battery-powered pacemakers died following device implantation compared to 1 of 10 (10%) mice implanted with wireless-powered pacemakers. Moreover, device function was achieved for 30 days with the wireless-powered device compared to 6 days with the battery-powered device. The wireless-powered pacemaker system presented herein will allow electrophysiology studies in numerous genetically engineered mouse models as well as rapid pacing-induced heart failure and atrial arrhythmia in mice.

  18. A Fully Implantable Pacemaker for the Mouse: From Battery to Wireless Power

    Science.gov (United States)

    Zellmer, Erik R.; Weinheimer, Carla J.; MacEwan, Matthew R.; Cui, Sophia X.; Nerbonne, Jeanne M.; Efimov, Igor R.

    2013-01-01

    Animal models have become a popular platform for the investigation of the molecular and systemic mechanisms of pathological cardiovascular physiology. Chronic pacing studies with implantable pacemakers in large animals have led to useful models of heart failure and atrial fibrillation. Unfortunately, molecular and genetic studies in these large animal models are often prohibitively expensive or not available. Conversely, the mouse is an excellent species for studying molecular mechanisms of cardiovascular disease through genetic engineering. However, the large size of available pacemakers does not lend itself to chronic pacing in mice. Here, we present the design for a novel, fully implantable wireless-powered pacemaker for mice capable of long-term (>30 days) pacing. This design is compared to a traditional battery-powered pacemaker to demonstrate critical advantages achieved through wireless inductive power transfer and control. Battery-powered and wireless-powered pacemakers were fabricated from standard electronic components in our laboratory. Mice (n = 24) were implanted with endocardial, battery-powered devices (n = 14) and epicardial, wireless-powered devices (n = 10). Wireless-powered devices were associated with reduced implant mortality and more reliable device function compared to battery-powered devices. Eight of 14 (57.1%) mice implanted with battery-powered pacemakers died following device implantation compared to 1 of 10 (10%) mice implanted with wireless-powered pacemakers. Moreover, device function was achieved for 30 days with the wireless-powered device compared to 6 days with the battery-powered device. The wireless-powered pacemaker system presented herein will allow electrophysiology studies in numerous genetically engineered mouse models as well as rapid pacing-induced heart failure and atrial arrhythmia in mice. PMID:24194832

  19. A fully implantable pacemaker for the mouse: from battery to wireless power.

    Directory of Open Access Journals (Sweden)

    Jacob I Laughner

    Full Text Available Animal models have become a popular platform for the investigation of the molecular and systemic mechanisms of pathological cardiovascular physiology. Chronic pacing studies with implantable pacemakers in large animals have led to useful models of heart failure and atrial fibrillation. Unfortunately, molecular and genetic studies in these large animal models are often prohibitively expensive or not available. Conversely, the mouse is an excellent species for studying molecular mechanisms of cardiovascular disease through genetic engineering. However, the large size of available pacemakers does not lend itself to chronic pacing in mice. Here, we present the design for a novel, fully implantable wireless-powered pacemaker for mice capable of long-term (>30 days pacing. This design is compared to a traditional battery-powered pacemaker to demonstrate critical advantages achieved through wireless inductive power transfer and control. Battery-powered and wireless-powered pacemakers were fabricated from standard electronic components in our laboratory. Mice (n = 24 were implanted with endocardial, battery-powered devices (n = 14 and epicardial, wireless-powered devices (n = 10. Wireless-powered devices were associated with reduced implant mortality and more reliable device function compared to battery-powered devices. Eight of 14 (57.1% mice implanted with battery-powered pacemakers died following device implantation compared to 1 of 10 (10% mice implanted with wireless-powered pacemakers. Moreover, device function was achieved for 30 days with the wireless-powered device compared to 6 days with the battery-powered device. The wireless-powered pacemaker system presented herein will allow electrophysiology studies in numerous genetically engineered mouse models as well as rapid pacing-induced heart failure and atrial arrhythmia in mice.

  20. Radiotherapy for breast cancer is not associated with increased risk of cied implantation

    DEFF Research Database (Denmark)

    Johansen, J. B.; Rehammar, J. C.; Jorgensen, O. D.

    2015-01-01

    Introduction: Radiotherapy is an important treatment in early stage breast cancer but it is claimed that radiotherapy causes damage to the cardiac conduction system and increases the risk implantation of CIED (pacemaker or ICD). However, this paradigm is based on smaller series of case reports. Due...... to the anatomy, radiotherapy will potential mainly affect the conduction system in left sided breast cancer. The aim of this study was to evaluate risk of implantation of a CIED subsequent to radiotherapy for breast cancer by comparing left- versus right sided radiotherapy in a nationwide cohort. Methods: From...... the database of the Danish Breast Cancer Collaborative Group, we identified women treated with radiotherapy for early-stage breast cancer in Denmark from 1982 to 2005. By record linkage to the Danish Pacemaker and ICD Registry information was retrieved on CIED implants subsequent to radiotherapy. The rate...

  1. End-to-side neurorrhaphy with and without perineurium

    Directory of Open Access Journals (Sweden)

    Fausto Viterbo

    Full Text Available OBJECTIVE: We compared end-to-side neurorraphy with and without the perineural sheath. METHOD: Twenty rats were used. The peroneal nerve was sectioned and the distal end was sutured to the lateral face of the tibial nerve. We removed the perineural sheath only on the right side, but not on the left side. The proximal end of the peroneal nerve was curved back approximately at a 100° angle and implanted into the adductor muscle. Six months later, the 14 surviving animals were submitted to electrophysiological tests, sacrificed, and the nerves and muscles were taken for histological exams. RESULTS: On the right side, the muscles that had positive response needed an average of 258.89 mV (±92.31 of electric stimulus and on the left side 298.34 mV (±139.32. The average weight of the tibial cranial muscles of the right side was 0.47 g (0.18 and for the left side 0.45 g (0.15. The distal end of the peroneal nerve showed averages of 310.29 (±191.34 nerve fibers on the right side and 287.71 (±183.60 on the left side. The tibial nerve above the neurorraphy showed averages of 939.46 (±223.51 nerve fibers on the right side and 959.46 (±327.48 on the left side. The tibial nerve below the neurorraphy showed averages of 935.17 (±298.65 nerve fibers on the right side and 755.31 (±323.26 on the left side. The average areas of the right tibial cranial muscles were 0.0162 m2 (±0.008, after 230 magnification, and 0.0152 m2 (0.0064 for the left tibial cranial muscles. The histological features of the tibial cranial muscles, taking normal as 100%, were 78.21 (±20.75 on the right side and 82.14 (±15.89 on the left side. The statistical analysis (Student's t test did not reveal any difference (p>0.05 among right and left sides for all variables. CONCLUSION: The authors concluded that the two neurorraphies (with and without perineurium did not show any difference regarding morphological and electrophysiological features studied.

  2. End-to-side neurorrhaphy with and without perineurium.

    Science.gov (United States)

    Viterbo, F; Teixeira, E; Hoshino, K; Padovani, C R

    1998-01-01

    We compared end-to-side neurorraphy with and without the perineural sheath. Twenty rats were used. The peroneal nerve was sectioned and the distal end was sutured to the lateral face of the tibial nerve. We removed the perineural sheath only on the right side, but not on the left side. The proximal end of the peroneal nerve was curved back approximately at a 100 degrees angle and implanted into the adductor muscle. Six months later, the 14 surviving animals were submitted to electrophysiological tests, sacrificed, and the nerves and muscles were taken for histological exams. On the right side, the muscles that had positive response needed an average of 258.89 mV (+/- 92.31) of electric stimulus and on the left side 298.34 mV (+/- 139.32). The average weight of the tibial cranial muscles of the right side was 0.47 g (0.18) and for the left side 0.45 g (0.15). The distal end of the peroneal nerve showed averages of 310.29 (+/- 191.34) nerve fibers on the right side and 287.71 (+/- 183.60) on the left side. The tibial nerve above the neurorraphy showed averages of 939.46 (+/- 223.51) nerve fibers on the right side and 959.46 (+/- 327.48) on the left side. The tibial nerve below the neurorraphy showed averages of 935.17 (+/- 298.65) nerve fibers on the right side and 755.31 (+/- 323.26) on the left side. The average areas of the right tibial cranial muscles were 0.0162 m2 (+/- 0.008), after 230 magnification, and 0.0152 m2 (0.0064) for the left tibial cranial muscles. The histological features of the tibial cranial muscles, taking normal as 100%, were 78.21 (+/- 20.75) on the right side and 82.14 (+/- 15.89) on the left side. The statistical analysis (Student's t test) did not reveal any difference (p > 0.05) among right and left sides for all variables. The authors concluded that the two neurorraphies (with and without perineurium) did not show any difference regarding morphological and electrophysiological features studied.

  3. Insights into Intrinsic Brain Networks based on Graph Theory and PET in right- compared to left-sided Temporal Lobe Epilepsy

    OpenAIRE

    Thomas Vanicek; Andreas Hahn; Tatjana Traub-Weidinger; Eva Hilger; Marie Spies; Wolfgang Wadsak; Rupert Lanzenberger; Ekaterina Pataraia; Susanne Asenbaum-Nan

    2016-01-01

    The human brain exhibits marked hemispheric differences, though it is not fully understood to what extent lateralization of the epileptic focus is relevant. Preoperative [18F]FDG-PET depicts lateralization of seizure focus in patients with temporal lobe epilepsy and reveals dysfunctional metabolic brain connectivity. The aim of the present study was to compare metabolic connectivity, inferred from inter-regional [18F]FDG PET uptake correlations, in right-sided (RTLE; n?=?30) and left-sided TL...

  4. Permanent pacemaker lead induced severe tricuspid regurgitation in patient undergoing multiple valve surgery.

    Science.gov (United States)

    Lee, Jung Hee; Kim, Tae Ho; Kim, Wook Sung

    2015-04-01

    Severe and permanent tricuspid regurgitation induced by pacemaker leads is rarely reported in the literature. The mechanism of pacemaker-induced tricuspid regurgitation has been identified, but its management has not been well established. Furthermore, debate still exists regarding the proper surgical approach. We present the case of a patient with severe tricuspid regurgitation induced by a pacemaker lead, accompanied by triple valve disease. The patient underwent double valve replacement and tricuspid valve repair without removal of the pre-existing pacemaker lead. The operation was successful and the surgical procedure is discussed in detail.

  5. Twiddler-syndrom er en årsag til pacemaker elektrode displacering

    DEFF Research Database (Denmark)

    Grønbech, Keea Treu; Hansen, Michael Gilså

    2013-01-01

    Twiddler's syndrome is a rare cause of pacemaker electrode displacement. The displacement is caused by the patient's manipulation with the pacemaker, so the electrode is retracted. We describe a case of a 79-year-old overweight woman with a known psychiatric anamnesis, who was admitted twice...

  6. Pacemaker lead perforation of the right ventricle associated with Moraxella phenylpyruvica infection in a dog.

    Science.gov (United States)

    Ciavarella, A; Nimmo, J; Hambrook, L

    2016-04-01

    A 13-year-old neutered male Border Collie was presented with acute onset syncope, weakness and anorexia 10 months after transvenous pacemaker implantation. The patient was laterally recumbent, bradycardic (36 beats/min) and febrile (40.7°C) on presentation. An electrocardiogram (ECG) revealed recurrence of third-degree atrioventricular block with a ventricular escape rhythm. Fluoroscopy identified migration of the pacemaker tip through the apex of the right ventricle. Echocardiography failed to reveal any evidence of pericardial effusion or cardiac tamponade. Full postmortem was performed after euthanasia. The pacemaker lead had perforated the apex of the right ventricle and lodged in the right pleural space. Culture of blood (taken antemortem), pericardial sac, right ventricular wall (surrounding pacemaker lead), pacemaker lead tip and pericardial fluid revealed a pure growth of Moraxella phenylpyruvica. Bacteraemia associated with M. phenylpyruvica has never been reported in the dog, but sporadic cases are reported in humans. Infection could have resulted from either pre-existing myocarditis or opportunistic infection and bacteraemia post pacemaker implantation. Evaluation of the pacemaker function at regular intervals would allow early detection of poor pacemaker-to-myocardium contact, which would prompt further investigation of pacemaker lead abnormalities such as perforation. © 2016 Australian Veterinary Association.

  7. Leadless cardiac pacemakers: present and the future.

    Science.gov (United States)

    Chew, Derek S; Kuriachan, Vikas

    2018-01-01

    Pacing technology for many decades has been composed of a generator attached to leads that are usually transvenous. Recently, leadless pacemakers have been studied in clinical settings and now available for use in many countries. This includes the single-component Nanostim Leadless Cardiac Pacemaker and Micra Transcatheter Pacing System, as well as the multicomponent Wireless Stimulation Endocardial system. Clinical studies in single-component leadless pacing technology has shown that they can be successfully implanted with minimal complications. The follow-up studies also seem to confirm the findings from the initial clinical trials. These systems offer some advantages over a traditional pacing system comprised of a subcutaneous generator and transvenous leads. In many ways, these leadless systems are disruptive technologies that are changing the traditional pacemaker concept and preferred for some patients. Ongoing research is needed to better assess their long-term function, safety, and end-of-life strategies. In the future, multichamber leadless pacing is expected to be developed and perhaps obviating the need for transvenous leads and their associated complications.

  8. Missed diagnosis of atresia of the right pulmonary artery in woman with left-sided pneumothorax

    DEFF Research Database (Denmark)

    Dagnegård, Hanna; Ryom, Philip

    2016-01-01

    Isolated pulmonary atresia is an uncommon condition, which can go undiagnosed for a long time in asymptomatic patients. Sometimes, diagnosis can be made at pregnancy due to respiratory symptoms. There is no known increased risk of pneumothorax. We here present a case where a second-time pregnant...... woman with an unknown atresia of the right pulmonary artery received a left-sided pneumothorax. The diagnosis was initially missed in spite of adequate imaging and the condition progressed to respiratory stop. We describe the course of diagnostics and the chosen strategy of treatment....

  9. Successful management of multiple permanent pacemaker complications – infection, 13 year old silent lead perforation and exteriorisation following failed percutaneous extraction, superior vena cava obstruction, tricuspid valve endocarditis, pulmonary embolism and prosthetic tricuspid valve thrombosis

    Science.gov (United States)

    Kaul, Pankaj; Adluri, Krishna; Javangula, Kalyana; Baig, Wasir

    2009-01-01

    A 59 year old man underwent mechanical tricuspid valve replacement and removal of pacemaker generator along with 4 pacemaker leads for pacemaker endocarditis and superior vena cava obstruction after an earlier percutaneous extraction had to be abandoned, 13 years ago, due to cardiac arrest, accompanied by silent, unsuspected right atrial perforation and exteriorisation of lead. Postoperative course was complicated by tricuspid valve thrombosis and secondary pulmonary embolism requiring TPA thrombolysis which was instantly successful. A review of literature of pacemaker endocarditis and tricuspid thrombosis along with the relevant management strategies is presented. We believe this case report is unusual on account of non operative management of right atrial lead perforation following an unsuccessful attempt at percutaneous removal of right sided infected pacemaker leads and the incidental discovery of the perforated lead 13 years later at sternotomy, presentation of pacemaker endocarditis with a massive load of vegetations along the entire pacemaker lead tract in superior vena cava, right atrial endocardium, tricuspid valve and right ventricular endocardium, leading to a functional and structural SVC obstruction, requirement of an unusually large dose of warfarin postoperatively occasioned, in all probability, by antibiotic drug interactions, presentation of tricuspid prosthetic valve thrombosis uniquely as vasovagal syncope and isolated hypoxia and near instantaneous resolution of tricuspid prosthetic valve thrombosis with Alteplase thrombolysis. PMID:19239701

  10. Successful management of multiple permanent pacemaker complications – infection, 13 year old silent lead perforation and exteriorisation following failed percutaneous extraction, superior vena cava obstruction, tricuspid valve endocarditis, pulmonary embolism and prosthetic tricuspid valve thrombosis

    Directory of Open Access Journals (Sweden)

    Javangula Kalyana

    2009-02-01

    Full Text Available Abstract A 59 year old man underwent mechanical tricuspid valve replacement and removal of pacemaker generator along with 4 pacemaker leads for pacemaker endocarditis and superior vena cava obstruction after an earlier percutaneous extraction had to be abandoned, 13 years ago, due to cardiac arrest, accompanied by silent, unsuspected right atrial perforation and exteriorisation of lead. Postoperative course was complicated by tricuspid valve thrombosis and secondary pulmonary embolism requiring TPA thrombolysis which was instantly successful. A review of literature of pacemaker endocarditis and tricuspid thrombosis along with the relevant management strategies is presented. We believe this case report is unusual on account of non operative management of right atrial lead perforation following an unsuccessful attempt at percutaneous removal of right sided infected pacemaker leads and the incidental discovery of the perforated lead 13 years later at sternotomy, presentation of pacemaker endocarditis with a massive load of vegetations along the entire pacemaker lead tract in superior vena cava, right atrial endocardium, tricuspid valve and right ventricular endocardium, leading to a functional and structural SVC obstruction, requirement of an unusually large dose of warfarin postoperatively occasioned, in all probability, by antibiotic drug interactions, presentation of tricuspid prosthetic valve thrombosis uniquely as vasovagal syncope and isolated hypoxia and near instantaneous resolution of tricuspid prosthetic valve thrombosis with Alteplase thrombolysis.

  11. Cardiac pacemaker. [electric-battery powered

    Energy Technology Data Exchange (ETDEWEB)

    Kolenik, S A

    1976-01-02

    The construction of a cardiac pacemaker is described which is characterized by particularly small dimensions, small weight and long life duration. The weight is under 100g, the specific weight under 1.7. Mass inertia forces which occur through acceleration and retardation processes, thus remain below the threshold values, above which one would have to reckon with considerable damaging of the surrounding body tissue. The maintaining of small size and slight weight is achieved by using an oscillator on COSMOS basis, where by considerably lower energy consumption, among others the lifetimes of the batteries used - a lithium anode with thionyl chloride electrolyte - is extended to over 5 years. The reliability can be increased by the use of 2 or more batteries. The designed dimension are 20x60x60 mm/sup 3/.

  12. Permanent Pacemaker Lead Induced Severe Tricuspid Regurgitation in Patient Undergoing Multiple Valve Surgery

    Directory of Open Access Journals (Sweden)

    Jung Hee Lee

    2015-04-01

    Full Text Available Severe and permanent tricuspid regurgitation induced by pacemaker leads is rarely reported in the literature. The mechanism of pacemaker-induced tricuspid regurgitation has been identified, but its management has not been well established. Furthermore, debate still exists regarding the proper surgical approach. We present the case of a patient with severe tricuspid regurgitation induced by a pacemaker lead, accompanied by triple valve disease. The patient underwent double valve replacement and tricuspid valve repair without removal of the pre-existing pacemaker lead. The operation was successful and the surgical procedure is discussed in detail.

  13. Implantable Cardiac Pacemakers – 50 Years from the First Implantation

    Directory of Open Access Journals (Sweden)

    Ratko Magjarević

    2010-01-01

    Overview: Development of implantable cardiac pacemaker was enabled by another important invention, the silicon transistor. h ough the invention of suitable lithium cells as appropriate power supply was essential for prolongation of battery life cycle and for increased reliability of pacemakers, main milestones in the development were associated with technological breakthroughs in electronics: from transistors, which introduced such features as small size and low power consumption, to hybrid and integrated circuits, which enabled programmability, microprocessors, which added more options in programming (multiprogrammability, diagnostics and telemetry, and the ICT (information communication technology that enabled physicians remote access to patients and interrogation of their implantable devices. Conclusions: Implantable pacemakers are reliable devices indicated for a wide range of dif erent therapies of cardiac rhythm disorders and heart failure. h ere is still a lot to learn about the physiology of a normal heart and even more about the failing heart. Modern pacemakers provide physicians valuable information from pacemakers’ memory via the built-in telemetry system. h ese information help physicians to better understand pathologic processes within the heart, thus contributing to the development of new ideas for treatment of diseases and for precise tailoring of the therapy to the patient’s needs. Although implantable pacemakers have reached the level of mature technology, they will continue to develop with therapies and diagnostics to facilitate a higher quality of life.

  14. [Left- or right-handed: the effect of a preferential use of one hand or the other on dental hygiene].

    Science.gov (United States)

    Eleveld, C A; Schuller, A A

    2016-02-01

    A research project investigated the extent to which a preferential use of one hand or the other has an effect on dental hygiene on the left or right side of the mouth. The study made use of epidemiological dental-care data from the Netherlands Organisation for Applied Scientific Research and of data from a dental practice specifically collected for this project. The results revealed that among a population which is 85-90% right-handed, statistically significantly more dental plaque was found on the right side of the mouth than on the left. A separate study revealed the prevalence of statistically significantly more dental plaque on the right side than on the left among right-handed people and, among left-handed people, a non-statistically significant trend of more dental plaque on the left than the right. It is concluded that dental hygiene on the left side and the right side of the mouth is very likely to be dependent on the preferential use of one hand or the other. The differences between the left side of the mouth and right among left- and right-handed people are, however, so small that it is questionable whether these should be taken into consideration in giving instructions about dental hygiene.

  15. Giant left atrium encountered during right-sided thoracentesis

    Directory of Open Access Journals (Sweden)

    Rashmi Advani

    Full Text Available Giant left atrium is an uncommon pathology to encounter during bedside chest ultrasound, but is an important structure to be aware of when considering thoracentesis. This cardiac structure could easily be mistaken for loculated pleural fluid. This case also supports growing evidence that expert users can safely perform thoracentesis without completely reversing therapeutic anticoagulation. Keywords: Giant left atrium, Thoracentesis, Rheumatic heart disease, Ultrasound

  16. Body-specific representations of action word meanings in right and left handers

    OpenAIRE

    Daniel Casasanto

    2007-01-01

    If understanding action words involves mentally simulating our own actions, then the neurocognitive representation of word meanings must differ for people with different kinds of bodies, who perform actions in systematically different ways. In a test of the _Body-Specificity Hypothesis_, right- and left-handers were compared on two motor-meaning congruity tasks. Double dissociations in both action execution and recognition memory results showed that right and left handers form body-specific r...

  17. A proposal for a standard communication protocol for pacemaker/ICD programmers

    NARCIS (Netherlands)

    Dijk, WA; Hooijschuur, CAM; van der Velde, W; Dassen, WRM

    2005-01-01

    The information generated by pacemakers and ICD's to support the cardiologist and technician for installing the optimal settings for the patient is increasing rapidly. In this paper a proposal is described for electronic data exchange between the pacemaker/ICD programmers and electronic information

  18. Circadian Pacemaker – Temperature Compensation

    NARCIS (Netherlands)

    Gerkema, Menno P.; Binder, Marc D.; Hirokawa, Nobutaka; Windhorst, Uwe

    2009-01-01

    One of the defining characteristics of circadian pacemakers and indicates the independence of the speed of circadian clock processes of environmental temperature. Mechanisms involved, so far not elucidated in full detail, entail at least two processes that are similarly affected by temperature

  19. Development of pacemaker properties and rhythmogenic mechanisms in the mouse embryonic respiratory network

    Science.gov (United States)

    Chevalier, Marc; Toporikova, Natalia; Simmers, John; Thoby-Brisson, Muriel

    2016-01-01

    Breathing is a vital rhythmic behavior generated by hindbrain neuronal circuitry, including the preBötzinger complex network (preBötC) that controls inspiration. The emergence of preBötC network activity during prenatal development has been described, but little is known regarding inspiratory neurons expressing pacemaker properties at embryonic stages. Here, we combined calcium imaging and electrophysiological recordings in mouse embryo brainstem slices together with computational modeling to reveal the existence of heterogeneous pacemaker oscillatory properties relying on distinct combinations of burst-generating INaP and ICAN conductances. The respective proportion of the different inspiratory pacemaker subtypes changes during prenatal development. Concomitantly, network rhythmogenesis switches from a purely INaP/ICAN-dependent mechanism at E16.5 to a combined pacemaker/network-driven process at E18.5. Our results provide the first description of pacemaker bursting properties in embryonic preBötC neurons and indicate that network rhythmogenesis undergoes important changes during prenatal development through alterations in both circuit properties and the biophysical characteristics of pacemaker neurons. DOI: http://dx.doi.org/10.7554/eLife.16125.001 PMID:27434668

  20. Magnetic Resonance Imaging in a Patient with a Dual Chamber Pacemaker

    Directory of Open Access Journals (Sweden)

    Lynne Martina Millar

    2010-01-01

    Full Text Available Having a pacemaker has been seen an absolute contraindication to having an MRI scan. This has become increasingly difficult in clinical practice as insertion of pacemakers and implantable cardiac defibrillators is at an all time high. Here we outline a case where a 71-year-old male patient with a permanent pacemaker needed to have an MRI scan to ascertain the aetiology of his condition and help guide further management. Given this clinical dilemma, an emergency clinical ethics consultation was arranged. As a result the patient underwent an MRI scan safely under controlled conditions with a consultant cardiologist and radiologist present. The results of the MRI scan were then able to tailor further treatment. This case highlights that in certain conditions an MRI can be performed in patients with permanent pacemakers and outlines the role of clinical ethics committees in complex medical decision making.

  1. Zebrafish heart as a model to study the integrative autonomic control of pacemaker function

    Science.gov (United States)

    Stoyek, Matthew R.; Quinn, T. Alexander; Croll, Roger P.

    2016-01-01

    The cardiac pacemaker sets the heart's primary rate, with pacemaker discharge controlled by the autonomic nervous system through intracardiac ganglia. A fundamental issue in understanding the relationship between neural activity and cardiac chronotropy is the identification of neuronal populations that control pacemaker cells. To date, most studies of neurocardiac control have been done in mammalian species, where neurons are embedded in and distributed throughout the heart, so they are largely inaccessible for whole-organ, integrative studies. Here, we establish the isolated, innervated zebrafish heart as a novel alternative model for studies of autonomic control of heart rate. Stimulation of individual cardiac vagosympathetic nerve trunks evoked bradycardia (parasympathetic activation) and tachycardia (sympathetic activation). Simultaneous stimulation of both vagosympathetic nerve trunks evoked a summative effect. Effects of nerve stimulation were mimicked by direct application of cholinergic and adrenergic agents. Optical mapping of electrical activity confirmed the sinoatrial region as the site of origin of normal pacemaker activity and identified a secondary pacemaker in the atrioventricular region. Strong vagosympathetic nerve stimulation resulted in a shift in the origin of initial excitation from the sinoatrial pacemaker to the atrioventricular pacemaker. Putative pacemaker cells in the sinoatrial and atrioventricular regions expressed adrenergic β2 and cholinergic muscarinic type 2 receptors. Collectively, we have demonstrated that the zebrafish heart contains the accepted hallmarks of vertebrate cardiac control, establishing this preparation as a viable model for studies of integrative physiological control of cardiac function by intracardiac neurons. PMID:27342878

  2. Noise-induced effects on multicellular biopacemaker spontaneous activity: Differences between weak and strong pacemaker cells

    Science.gov (United States)

    Aghighi, Alireza; Comtois, Philippe

    2017-09-01

    Self-organization of spontaneous activity of a network of active elements is important to the general theory of reaction-diffusion systems as well as for pacemaking activity to initiate beating of the heart. Monolayer cultures of neonatal rat ventricular myocytes, consisting of resting and pacemaker cells, exhibit spontaneous activation of their electrical activity. Similarly, one proposed approach to the development of biopacemakers as an alternative to electronic pacemakers for cardiac therapy is based on heterogeneous cardiac cells with resting and spontaneously beating phenotypes. However, the combined effect of pacemaker characteristics, density, and spatial distribution of the pacemaker cells on spontaneous activity is unknown. Using a simple stochastic pattern formation algorithm, we previously showed a clear nonlinear dependency of spontaneous activity (occurrence and amplitude of spontaneous period) on the spatial patterns of pacemaker cells. In this study, we show that this behavior is dependent on the pacemaker cell characteristics, with weaker pacemaker cells requiring higher density and larger clusters to sustain multicellular activity. These multicellular structures also demonstrated an increased sensitivity to voltage noise that favored spontaneous activity at lower density while increasing temporal variation in the period of activity. This information will help researchers overcome the current limitations of biopacemakers.

  3. Devices That May Interfere with Pacemakers

    Science.gov (United States)

    ... Communications Commission (FCC) makes new frequencies available. Newer cellphones using these new frequencies might make pacemakers less reliable. A group of cellphone companies is studying that possibility. Bluetooth® headsets do ...

  4. Regulation of basal and reserve cardiac pacemaker function by interactions of cAMP mediated PKA-dependent Ca2+ cycling with surface membrane channels

    Science.gov (United States)

    Vinogradova, Tatiana M.; Lakatta, Edward G.

    2009-01-01

    Decades of intensive research of primary cardiac pacemaker, the sinoatrial node, have established potential roles of specific membrane channels in the generation of the diastolic depolarization, the major mechanism allowing sinoatrial node cells generate spontaneous beating. During the last three decades, multiple studies made either in the isolated sinoatrial node or sinoatrial node cells have demonstrated a pivotal role of Ca2+ and, specifically Ca2+-release from sarcoplasmic reticulum, for spontaneous beating of cardiac pacemaker. Recently, spontaneous, rhythmic local subsarcolemmal Ca2+ releases from ryanodine receptors during late half of the diastolic depolarization have been implicated as a vital factor in the generation of sinoatrial node cells spontaneous firing. Local Ca2+ releases are driven by a unique combination of high basal cAMP production by adenylyl cyclases, high basal cAMP degradation by phosphodiesterases and a high level of cAMP-mediated PKA-dependent phosphorylation. These local Ca2+ releases activate an inward Na+-Ca2+ exchange current which accelerates the terminal diastolic depolarization rate and, thus, controls the spontaneous pacemaker firing. Both the basal primary pacemaker beating rate and its modulation via β-adrenergic receptor stimulation appear to be critically dependent upon intact RyR function and local subsarcolemmal sarcoplasmic reticulum generated Ca2+ releases. This review aspires to integrate the traditional viewpoint that has emphasized the supremacy of the ensemble of surface membrane ion channels in spontaneous firing of the primary cardiac pacemaker, and these novel perspectives of cAMP-mediated PKA-dependent Ca2+ cycling in regulation of the heart pacemaker clock, both in the basal state and during β-adrenergic receptor stimulation. PMID:19573534

  5. Early Performance and Safety of the Micra Transcatheter Pacemaker in Pigs.

    Science.gov (United States)

    Bonner, Matthew; Eggen, Michael; Haddad, Tarek; Sheldon, Todd; Williams, Eric

    2015-11-01

    The Micra® Transcatheter Pacing System (TPS; Medtronic Inc., Minneapolis, MN, USA) is a miniaturized single-chamber pacemaker system that is delivered via catheter through the femoral vein. In this study, the electrical performance was compared between the TPS and a traditional leaded pacemaker. In addition, the safety profile of the two systems was compared by thorough monitoring for a number of adverse events. The TPS was implanted in the right ventricular apex of 10 Yucatan mini pigs and a Medtronic single-lead pacemaker (SLP) was implanted in the right ventricular apex of another 10 pigs and connected to a traditional pacemaker. The electrical performance of all devices was monitored for 12 weeks. The safety profile of each system was characterized using x-ray, computed tomography, ultrasound, blood work, and necropsy to monitor for a variety of adverse events. At implant the mean pacing thresholds were 0.58 ± 0.17 V @0.24 ms and 0.75 ± 0.29 V @0.21 ms for the TPS and the SLP respectively. After 12 weeks, mean thresholds were 0.94 ± 0.46 V and 1.85 ± 0.75 V (P pacemaker system. © 2015 Medtronic PLC. Pacing and Clinical Electrophysiology published by Wiley Periodicals, Inc.

  6. Dragoljub (Bata Adamov (1927-1996: The first pacemaker implantation in Serbia

    Directory of Open Access Journals (Sweden)

    Pavlović Siniša U.

    2017-01-01

    Full Text Available It has been over half a century since the implementation of pacemaker therapy in our country and the region. The first successful implantation of a pacemaker in former Yugoslavia and in Serbia took place on September 16, 1965 in “Dr. Dragiša Mišović” Clinical Hospital Centre, and this operation, with a team of doctors of the institution, was performed by surgeon Dragoljub (Bata Adamov (1927–1996. The first permanent pacemaker implantation was with epicardial leads with thoracotomy approach. The patient was operated on under general anesthesia, administered by anesthesiologist Predrag Lalević (1927–, and Dr. Adamov was assisted by Dr. Miša Albrecht (1933– and Dr. Milan Dragović (1933–2009. Although pacemaker therapy has since been widely proven and confirmed, it is necessary to remember the pioneers who introduced this kind of therapy to the region, as they deserve a distinguished place in the history of medicine in Serbia.

  7. A leadless pacemaker in the real-world setting

    DEFF Research Database (Denmark)

    Roberts, Paul R.; Clementy, Nicolas; Al Samadi, Faisal

    2017-01-01

    , telemetry, and battery issues. Objective The acute performance of the Micra transcatheter pacemaker from a worldwide Post-Approval Registry is reported. Methods The registry is an ongoing prospective single-arm observational study designed to assess the safety and effectiveness of Micra in the post...... were low and stable. Conclusion Performance of the Micra transcatheter pacemaker in a real-world setting demonstrates a high rate (99.6%) of implant success and low rate (1.51%) of major complications through 30 days post implant. In particular, the rates of pericardial effusion, device dislodgement...

  8. Atrioventricular Pacemaker Lead Reversal

    Directory of Open Access Journals (Sweden)

    Mehmet K Aktas, MD

    2007-01-01

    Full Text Available During cardiac surgery temporary epicardial atrial and ventricular leads are placed in case cardiac pacing is required postoperatively. We present the first reported series of patients with reversal of atrioventricular electrodes in the temporary pacemaker without any consequent deleterious hemodynamic effect. We review the electrocardiographic findings and discuss the findings that lead to the discovery of atrioventricular lead reversal.

  9. Nanomedicinal products: a survey on specific toxicity and side effects

    Directory of Open Access Journals (Sweden)

    Brand W

    2017-08-01

    Full Text Available Walter Brand,1,* Cornelle W Noorlander,1,* Christina Giannakou,2,3 Wim H De Jong,2 Myrna W Kooi,1 Margriet VDZ Park,2 Rob J Vandebriel,2 Irene EM Bosselaers,4 Joep HG Scholl,5 Robert E Geertsma2 1Centre for Safety of Substances and Products, 2Centre for Health Protection, National Institute for Public Health and the Environment (RIVM, Bilthoven, 3Department of Toxicogenomics, Maastricht University, Maastricht, 4Section Pharmacology, Toxicology and Pharmacokinetics, Medicines Evaluation Board (CBG-MEB, Utrecht, 5Research & Analysis Department, Netherlands Pharmacovigilance Centre Lareb, ‘s-Hertogenbosch, the Netherlands *These authors contributed equally to this work Abstract: Due to their specific properties and pharmacokinetics, nanomedicinal products (NMPs may present different toxicity and side effects compared to non-nanoformulated, conventional medicines. To facilitate the safety assessment of NMPs, we aimed to gain insight into toxic effects specific for NMPs by systematically analyzing the available toxicity data on approved NMPs in the European Union. In addition, by comparing five sets of products with the same active pharmaceutical ingredient (API in a conventional formulation versus a nanoformulation, we aimed to identify any side effects specific for the nano aspect of NMPs. The objective was to investigate whether specific toxicity could be related to certain structural types of NMPs and whether a nanoformulation of an API altered the nature of side effects of the product in humans compared to a conventional formulation. The survey of toxicity data did not reveal nanospecific toxicity that could be related to certain types of structures of NMPs, other than those reported previously in relation to accumulation of iron nanoparticles (NPs. However, given the limited data for some of the product groups or toxicological end points in the analysis, conclusions with regard to (a lack of potential nanomedicine-specific effects need to be

  10. Dual chamber pacemaker implants - a new opportunity in Pakistan for children with congenital and acquired complete heart block

    International Nuclear Information System (INIS)

    Ashfaq, A.; Khan, M.A.; Atiq, M.

    2011-01-01

    Implantation of cardiac pacemakers has been practiced for at least five decades with continuous developments of the hardware. The invention of dual chamber pacemakers has initiated a debate concerning its superiority over single chamber ventricular pacemakers. Throughout the world, surgeons have been using dual chambered permanent pacemakers with successful follow ups. However, Pakistan has not yet taken the advantage of such pacemaker devices till now. We report three cases that underwent a dual chamber permanent pacemaker implantation for the first time in children less than 8 kg with successful follow ups. (author)

  11. Model-based approach for quantitative estimates of skin, heart, and lung toxicity risk for left-side photon and proton irradiation after breast-conserving surgery.

    Science.gov (United States)

    Tommasino, Francesco; Durante, Marco; D'Avino, Vittoria; Liuzzi, Raffaele; Conson, Manuel; Farace, Paolo; Palma, Giuseppe; Schwarz, Marco; Cella, Laura; Pacelli, Roberto

    2017-05-01

    Proton beam therapy represents a promising modality for left-side breast cancer (BC) treatment, but concerns have been raised about skin toxicity and poor cosmesis. The aim of this study is to apply skin normal tissue complication probability (NTCP) model for intensity modulated proton therapy (IMPT) optimization in left-side BC. Ten left-side BC patients undergoing photon irradiation after breast-conserving surgery were randomly selected from our clinical database. Intensity modulated photon (IMRT) and IMPT plans were calculated with iso-tumor-coverage criteria and according to RTOG 1005 guidelines. Proton plans were computed with and without skin optimization. Published NTCP models were employed to estimate the risk of different toxicity endpoints for skin, lung, heart and its substructures. Acute skin NTCP evaluation suggests a lower toxicity level with IMPT compared to IMRT when the skin is included in proton optimization strategy (0.1% versus 1.7%, p < 0.001). Dosimetric results show that, with the same level of tumor coverage, IMPT attains significant heart and lung dose sparing compared with IMRT. By NTCP model-based analysis, an overall reduction in the cardiopulmonary toxicity risk prediction can be observed for all IMPT compared to IMRT plans: the relative risk reduction from protons varies between 0.1 and 0.7 depending on the considered toxicity endpoint. Our analysis suggests that IMPT might be safely applied without increasing the risk of severe acute radiation induced skin toxicity. The quantitative risk estimates also support the potential clinical benefits of IMPT for left-side BC irradiation due to lower risk of cardiac and pulmonary morbidity. The applied approach might be relevant on the long term for the setup of cost-effectiveness evaluation strategies based on NTCP predictions.

  12. A Functional Analysis of Circadian Pacemakers in Nocturnal Rodents. IV. Entrainment : Pacemaker as Clock

    NARCIS (Netherlands)

    Pittendrigh, Colin S.; Daan, Serge

    1976-01-01

    1. In the first part of the paper, the model of non-parametric entrainment of circadian pacemakers is tested for the case of nocturnal rodents. The model makes use of the available data on freerunning period (τ) in constant darkness and on phase response curves (PRC) for short light pulses. It is

  13. Sikkerhed af magnetisk resonans-skanning hos patienter med pacemaker og implanterbar defibrillator

    DEFF Research Database (Denmark)

    Al-Sabagh, Kifah Hekmat; Christensen, Britta Ege; Thøgersen, Anna Margrethe

    2010-01-01

    INTRODUCTION: The presence of a cardiac implantable device is ICD considered an absolute contraindication to magnetic resonance imaging (MRI). The purpose of this study was to evaluate the safety of performing MRI in patients with cardiac pacemakers and ICDs that had a compelling clinical need...... for MRI examination. MATERIAL AND METHODS: During a period of nine years we have included 65 patients with cardiac devices (60 pacemakers and five ICDs) who underwent a total of 73 MRI examinations at 1.5 T. All pacemakers were reprogrammed before MRI to asynchronous mode to avoid MRI-induced inhibition...... safely in 63 patients. Inhibition of pacemaker output was observed in one patient and induction of ventricular fibrillation was observed in another with ICD. A significant increase in PCT was rare and only detected in 1% of all electrodes. CONCLUSION: MRI can be performed safely in patients...

  14. Remote control improves quality of life in elderly pacemaker patients versus standard ambulatory-based follow-up.

    Science.gov (United States)

    Comoretto, Rosanna Irene; Facchin, Domenico; Ghidina, Marco; Proclemer, Alessandro; Gregori, Dario

    2017-08-01

    Health-related quality of life (HRQoL) improves shortly after pacemaker (PM) implantation. No studies have investigated the HRQoL trend for elderly patients with a remote device monitoring follow-up system. Using EuroQol-5D Questionnaire and the PM-specific Assessment of Quality of Life and Related Events Questionnaire, HRQoL was measured at baseline and then repeatedly during the 6 months following PM implantation in a cohort of 42 consecutive patients. Twenty-five patients were followed-up with standard outpatient visits, while 17 used a remote monitoring system. Aquarel scores were significantly higher in patients with remote device monitoring system regarding chest discomfort and arrhythmia subscales the first month after PM implant and remained stable until 6 months. Remote monitoring affected the rate of HRQoL improvement in the first 3 months after pacemaker implantation more than ambulatory follow-up. Remote device monitoring has a significant impact on HRQoL in pacemaker patients, increasing its levels up to 6 months after implant. © 2017 John Wiley & Sons, Ltd.

  15. Leadless Cardiac Pacemaker Implantation After Lead Extraction in Patients With Severe Device Infection.

    Science.gov (United States)

    Kypta, Alexander; Blessberger, Hermann; Kammler, Juergen; Lambert, Thomas; Lichtenauer, Michael; Brandstaetter, Walter; Gabriel, Michael; Steinwender, Clemens

    2016-09-01

    Conventional pacemaker therapy is limited by short- and long-term complications, most notably device infection. Transcatheter pacing systems (TPS) may be beneficial in this kind of patients as they eliminate the need for a device pocket and leads and thus may reduce the risk of re-infection. We assessed a novel procedure in 6 patients with severe device infection who were pacemaker dependent. After lead extraction a single chamber TPS was implanted into the right ventricle. Of the 6 patients who underwent lead extraction due to severe device infection at our institution, 3 were diagnosed with a pocket infection only, whereas the other 3 showed symptoms of both pocket and lead infection. Successful lead extraction and TPS implantation was accomplished in all patients. Four patients were bridged with a temporary pacemaker between 2 hours and 2 days after lead extraction, whereas 2 patients had the TPS implanted during the same procedure just before traditional pacemaker system removal. All patients stayed free of infection during the follow-up period of 12 weeks. An additional positron emission tomography scan was performed in each patient and indicated no signs of an infection around the TPS. Transcather pacemaker implantation was safe and feasible in 6 patients and did not result in re-infection even if implanted before removal of the infected pacemaker system within the same procedure. Therefore, implantation of a TPS may be an option for patients with severe device infection, especially in those with blocked venous access or who are pacemaker dependent. © 2016 Wiley Periodicals, Inc.

  16. Pacemakers in patients with familial dysautonomia--a review of experience with 20 patients.

    Science.gov (United States)

    Gold-von Simson, Gabrielle; Rutkowski, Monika; Berlin, Dena; Axelrod, Felicia B

    2005-02-01

    Familial dysautonomia (FD) is a genetic disease associated with a high incidence of sudden death. If fatal bradyarrhythmia is an etiological factor then the incidence of sudden death should decrease after pacemaker placement. Retrospective review of 596 registered FD patients revealed that 22 FD patients (3.7%) had pacemakers placed between December 1984 and June 2003. Clinical and electrocardiographic indications for placement and demographic data were assessed for 20 of the 22 patients (10 males, 10 females, ages 4 to 48 years). Two patients were excluded because of insufficient data. Prior to pacemaker placement, presenting symptoms were syncope and cardiac arrest, 16/20 (80%) and 6/20 (30 %), respectively. Asystole was the most frequent electrocardiographic finding and was documented in 17/20 patients (85 %). Other electrocardiographic abnormalities included bradycardia, AV block, prolonged QTc and prolonged JTc. The average duration of pacemaker utilization was 5.7 years (range 5 months to 14.5 years). Complications included infection (1 patient) and wire migration (2 patients). In the one patient with infection, the pacemaker was permanently removed. This patient then experienced multiple syncopal episodes and death. There were 7 other deaths. Three deaths occurred suddenly without preceding events, and 4 patients had non-cardiac causes of death. None of these 7 deceased patients had recurrence of syncope after pacemaker placement. In the 12 surviving patients, 6 had recurrence of syncope but none had cardiac arrest. Pacemaker placement may protect FD patients from fatal bradyarrhythmia and may decrease the incidence of syncope. However, data are limited and prospective analysis is needed.

  17. Elektrokirurgi hos patienter med pacemaker og implanterbar kardioverter-defibrillator

    DEFF Research Database (Denmark)

    Lønnberg, Ann Sophie Claire; Philberts, Berit Thornvig; Bonde, Christian

    2017-01-01

    Electrosurgery is a very useful tool and one of the most commonly used techniques. However, the technique can interfere with pacemakers and implantable cardioverter defibrillators. This article provides practical recommendations for the use of electrosurgery in these patients.......Electrosurgery is a very useful tool and one of the most commonly used techniques. However, the technique can interfere with pacemakers and implantable cardioverter defibrillators. This article provides practical recommendations for the use of electrosurgery in these patients....

  18. Isolated persistent left-sided superior vena cava, giant coronary sinus, atrial tachycardia and heart failure in a child

    Directory of Open Access Journals (Sweden)

    Nagaraja Moorthy

    2013-09-01

    Full Text Available Persistence of a left-sided superior vena cava (PLSVC with absent right superior vena cava (isolated PLSVC is a very rare venous malformation and commonly associated with congenital heart disease or alterations of the cardiac situs. We describe an unusual case of a young boy presenting with persistent atrial tachycardia and congestive heart failure. He was detected to have unexplained grossly dilated right atrium, right ventricle with systolic dysfunction and a giant coronary sinus (CS. The dilated CS closely mimicked a pseudo cor-triatriatum on echocardiography. Contrast echocardiography from both arms revealed opacification of the CS before the right atrium. Bilateral upper limb venography confirmed the presence of absent right SVC and isolated persistent left SVC draining into the giant coronary sinus.

  19. [Surgical therapy of life-threatening tachycardic cardiac arrhythmias in children].

    Science.gov (United States)

    Frank, G; Schmid, C; Baumgart, D; Lowes, D; Klein, H; Kallfelz, H C

    1989-05-01

    Surgical techniques for tachyarrhythmias refractory to medical treatment are used with increasing frequency. Among 211 patients undergoing antiarrhythmic surgery 10 children (2 to 14 years old) were operated by electrophysiologically directed procedures. 7 patients suffered from WPW syndrome, 2 from focal atrial tachycardias and 1 from recurrent ventricular tachycardia following the repair of Fallot's tetralogy. In all cases preoperative electrophysiologic study and intraoperative mapping preceded operative ablation. Surgical treatment consisted of interruption of the bundle of Kent (3 right-sided, 2 left-sided, 3 septal), ablation of the atrial focus (1 right-sided, 1 left-sided) and right ventricular outflow tract incision. In 7 operations cryo-techniques were added. 2 children with WPW syndrome had two interventions because of tachycardia recurrences due to multiple accessory pathways. In 1 case a VVI-pacemaker was implanted postoperatively due to complete atrioventricular block. Another 2 children with prolonged postoperative bradycardia received a pacemaker prophylactically. Only the child with previous tetralogy of Fallot is still under antiarrhythmic medication while all other children are free of tachycardiac episodes. Our data confirm the efficacy of surgical treatment of tachyarrhythmias in children thereby abolishing the need for life-long antiarrhythmic medication.

  20. Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

    Directory of Open Access Journals (Sweden)

    Taku Nishida, MD

    2009-01-01

    Full Text Available In May 2005 a 68-year-old woman received a VDD pacemaker implantation in the right pectoral region at our hospital for the treatment of complete atrioventricular block. In July 2008, she was diagnosed with dilated cardiomyopathy based on histological testing. In November 2008, she developed syncope due to ventricular tachycardia while at another hospital. She underwent external electrical cardioversion with an anterior-lateral paddle position using a single shock of 100 J. This shock led to severe bradycardia resulting in a transfer to our hospital. The physician who provided the shock could not have been aware that the patient had an implanted pacemaker. The skin above the pulse generator was burned. The electrocardiogram showed no pacing spikes or ventricular escape rhythm. Investigation of the pacemaker 3 hours after cardioversion revealed reprogramming of the device and a marked rise in the lead impedance (>3,000 ohm. Removal of the generator and implantation of a biventricular cardioverter defibrillator were required. The emergency situation, the small size of the generator, the small incision made using the buried suture method, and the patient's obesity all probably contributed to the physician's not noticing the implanted pacemaker. It is important to increase awareness of the severe consequences that may follow if the physician administering external defibrillation does not know about the patient's implanted pacemaker.

  1. Grammars with two-sided contexts

    Directory of Open Access Journals (Sweden)

    Mikhail Barash

    2014-05-01

    Full Text Available In a recent paper (M. Barash, A. Okhotin, "Defining contexts in context-free grammars", LATA 2012, the authors introduced an extension of the context-free grammars equipped with an operator for referring to the left context of the substring being defined. This paper proposes a more general model, in which context specifications may be two-sided, that is, both the left and the right contexts can be specified by the corresponding operators. The paper gives the definitions and establishes the basic theory of such grammars, leading to a normal form and a parsing algorithm working in time O(n^4, where n is the length of the input string.

  2. Circadian Activators Are Expressed Days before They Initiate Clock Function in Late Pacemaker Neurons from Drosophila.

    Science.gov (United States)

    Liu, Tianxin; Mahesh, Guruswamy; Houl, Jerry H; Hardin, Paul E

    2015-06-03

    Circadian pacemaker neurons in the Drosophila brain control daily rhythms in locomotor activity. These pacemaker neurons can be subdivided into early or late groups depending on whether rhythms in period (per) and timeless (tim) expression are initiated at the first instar (L1) larval stage or during metamorphosis, respectively. Because CLOCK-CYCLE (CLK-CYC) heterodimers initiate circadian oscillator function by activating per and tim transcription, a Clk-GFP transgene was used to mark when late pacemaker neurons begin to develop. We were surprised to see that CLK-GFP was already expressed in four of five clusters of late pacemaker neurons during the third instar (L3) larval stage. CLK-GFP is only detected in postmitotic neurons from L3 larvae, suggesting that these four late pacemaker neuron clusters are formed before the L3 larval stage. A GFP-cyc transgene was used to show that CYC, like CLK, is also expressed exclusively in pacemaker neurons from L3 larval brains, demonstrating that CLK-CYC is not sufficient to activate per and tim in late pacemaker neurons at the L3 larval stage. These results suggest that most late pacemaker neurons develop days before novel factors activate circadian oscillator function during metamorphosis. Copyright © 2015 the authors 0270-6474/15/358662-10$15.00/0.

  3. Rarity of invasiveness in right-sided infective endocarditis

    DEFF Research Database (Denmark)

    Hussain, Syed T; Shrestha, Nabin K; Witten, James

    2018-01-01

    OBJECTIVE: The rarity of invasiveness of right-sided infective endocarditis (IE) compared with left-sided has not been well recognized and evaluated. Thus, we compared invasiveness of right- versus left-sided IE in surgically treated patients. PATIENTS AND METHODS: From January 2002 to January 20...

  4. Biological pacemaker created by minimally invasive somatic reprogramming in pigs with complete heart block

    Science.gov (United States)

    Hu, Yu-Feng; Dawkins, James Frederick; Cho, Hee Cheol; Marbán, Eduardo; Cingolani, Eugenio

    2016-01-01

    Somatic reprogramming by reexpression of the embryonic transcription factor T-box 18 (TBX18) converts cardiomyocytes into pacemaker cells. We hypothesized that this could be a viable therapeutic avenue for pacemaker-dependent patients afflicted with device-related complications, and therefore tested whether adenoviral TBX18 gene transfer could create biological pacemaker activity in vivo in a large-animal model of complete heart block. Biological pacemaker activity, originating from the intramyocardial injection site, was evident in TBX18-transduced animals starting at day 2 and persisted for the duration of the study (14 days) with minimal backup electronic pacemaker use. Relative to controls transduced with a reporter gene, TBX18-transduced animals exhibited enhanced autonomic responses and physiologically superior chronotropic support of physical activity. Induced sinoatrial node cells could be identified by their distinctive morphology at the site of injection in TBX18-transduced animals, but not in controls. No local or systemic safety concerns arose. Thus, minimally invasive TBX18 gene transfer creates physiologically relevant pacemaker activity in complete heart block, providing evidence for therapeutic somatic reprogramming in a clinically relevant disease model. PMID:25031269

  5. Volumetric modulated arc therapy and breath-hold in image-guided locoregional left-sided breast irradiation

    International Nuclear Information System (INIS)

    Osman, Sarah O.S.; Hol, Sandra; Poortmans, Philip M.; Essers, Marion

    2014-01-01

    Purpose: To investigate the effects of using volumetric modulated arc therapy (VMAT) and/or voluntary moderate deep inspiration breath-hold (vmDIBH) in the radiation therapy (RT) of left-sided breast cancer including the regional lymph nodes. Materials and methods: For 13 patients, four treatment combinations were compared; 3D-conformal RT (i.e., forward IMRT) in free-breathing 3D-CRT(FB), 3D-CRT(vmDIBH), 2 partial arcs VMAT(FB), and VMAT(vmDIBH). Prescribed dose was 42.56 Gy in 16 fractions. For 10 additional patients, 3D-CRT and VMAT in vmDIBH only were also compared. Results: Dose conformity, PTV coverage, ipsilateral and total lung doses were significantly better for VMAT plans compared to 3D-CRT. Mean heart dose (D mean,heart ) reduction in 3D-CRT(vmDIBH) was between 0.9 and 8.6 Gy, depending on initial D mean,heart (in 3D-CRT(FB) plans). VMAT(vmDIBH) reduced the D mean,heart further when D mean,heart was still >3.2 Gy in 3D-CRT(vmDIBH). Mean contralateral breast dose was higher for VMAT plans (2.7 Gy) compared to 3DCRT plans (0.7 Gy). Conclusions: VMAT and 3D-CRT(vmDIBH) significantly reduced heart dose for patients treated with locoregional RT of left-sided breast cancer. When D mean,heart exceeded 3.2 Gy in 3D-CRT(vmDIBH) plans, VMAT(vmDIBH) resulted in a cumulative heart dose reduction. VMAT also provided better target coverage and reduced ipsilateral lung dose, at the expense of a small increase in the dose to the contralateral breast

  6. Quantitative assessment of left-sided valvular regurgitation using the fourier amplitude ratio

    International Nuclear Information System (INIS)

    Kosuda, Shigeru; Satoh, Jinsei; Yonahara, Yoshio; Asato, Tetsuyoshi; Naito, Masahito; Nishiguchi, Iku; Ogawa, Koichi; Kubo, Atsushi; Hashimoto, Shozo.

    1985-01-01

    Since the amplitude value of a pixel is proportional to the stroke counts of the pixel, the total amplitude value of each ventricle would reflect its own stroke volume. Stroke count ratio (SCR) and ventricular amplitude ratio (VAR) values were simultaneously calculated and compared in 43 subjects, including 13 subjects with valvular regurgitation, after multigated equilibrium scintigraphy was performed. Mean values of SCR in 13 subjects with valvular regurgitation and 30 control subjects were 2.22+-1.11, 1.24+-0.41, respectively (p<0.005). Mean values of VAR were 2.52+-0.87, 1.36+-0.39, respectively (p<0.0005). The VAR method was independent of the tilting angles of the detector, and showed excellent intra-observer and inter-observer reproducibilities (r=0.95, r=0.91). The VAR method derived from the Fourier amplitude image is a noninvasive technique, is suitable for serial studies, and appears to be a relatively reliable means of assessing the severity of left-sided valvular regurgitation. (author)

  7. Rarity of invasiveness in right-sided infective endocarditis.

    Science.gov (United States)

    Hussain, Syed T; Shrestha, Nabin K; Witten, James; Gordon, Steven M; Houghtaling, Penny L; Tingleff, Jens; Navia, José L; Blackstone, Eugene H; Pettersson, Gösta B

    2018-01-01

    The rarity of invasiveness of right-sided infective endocarditis (IE) compared with left-sided has not been well recognized and evaluated. Thus, we compared invasiveness of right- versus left-sided IE in surgically treated patients. From January 2002 to January 2015, 1292 patients underwent surgery for active IE, 138 right-sided and 1224 left-sided. Among patients with right-sided IE, 131 had tricuspid and 7 pulmonary valve IE; 12% had prosthetic valve endocarditis. Endocarditis-related invasiveness was based on echocardiographic and operative findings. Invasive disease was rare on the right side, occurring in 1 patient (0.72%; 95% confidence interval 0.02%-4.0%); rather, it was limited to valve cusps/leaflets or was superficial. In contrast, IE was invasive in 408 of 633 patients with aortic valve (AV) IE (65%), 113 of 369 with mitral valve (MV) IE (31%), and 148 of 222 with AV and MV IE (67%). Staphylococcus aureus was a more predominant organism in right-sided than left-sided IE (right 40%, AV 19%, MV 29%), yet invasion was observed almost exclusively on the left side of the heart, which was more common and more severe with AV than MV IE and more common with prosthetic valve endocarditis than native valve IE. Rarity of right-sided invasion even when caused by S aureus suggests that invasion and development of cavities/"abscesses" in patients with IE may be driven more by chamber pressure than organism, along with other reported host-microbial interactions. The lesser invasiveness of MV compared with AV IE suggests a similar mechanism: decompression of MV annulus invasion site(s) toward the left atrium. Copyright © 2017 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

  8. Automatic Capture Verification in Pacemakers (Autocapture – Utility and Problems

    Directory of Open Access Journals (Sweden)

    Ruth Kam

    2004-04-01

    Full Text Available The concept of a closed – loop feedback system, that would automatically assess pacing threshold and self -adjust pacing output to ensure consistent myocardial capture, has many appeals. Enhancing patient safety in cases of an unexpected rise in threshold, reduced current drain, hence prolonging battery longevity and reducing the amount of physician intervention required are just some of the advantages. Autocapture (AC is a proprietary algorithm developed by St Jude Medical CRMD, Sylmar, CA, USA, (SJM that was the first to commercially provide these automatic functions in a single chamber pacemaker (Microny and Regency, and subsequently in a dual chamber pacemaker (Affinity, Entity and Identity family of pacemakers. This article reviews the conditions necessary for AC verification and performance and the problems encountered in clinical practice.

  9. Insights into Intrinsic Brain Networks based on Graph Theory and PET in right- compared to left-sided Temporal Lobe Epilepsy.

    Science.gov (United States)

    Vanicek, Thomas; Hahn, Andreas; Traub-Weidinger, Tatjana; Hilger, Eva; Spies, Marie; Wadsak, Wolfgang; Lanzenberger, Rupert; Pataraia, Ekaterina; Asenbaum-Nan, Susanne

    2016-06-28

    The human brain exhibits marked hemispheric differences, though it is not fully understood to what extent lateralization of the epileptic focus is relevant. Preoperative [(18)F]FDG-PET depicts lateralization of seizure focus in patients with temporal lobe epilepsy and reveals dysfunctional metabolic brain connectivity. The aim of the present study was to compare metabolic connectivity, inferred from inter-regional [(18)F]FDG PET uptake correlations, in right-sided (RTLE; n = 30) and left-sided TLE (LTLE; n = 32) with healthy controls (HC; n = 31) using graph theory based network analysis. Comparing LTLE and RTLE and patient groups separately to HC, we observed higher lobar connectivity weights in RTLE compared to LTLE for connections of the temporal and the parietal lobe of the contralateral hemisphere (CH). Moreover, especially in RTLE compared to LTLE higher local efficiency were found in the temporal cortices and other brain regions of the CH. The results of this investigation implicate altered metabolic networks in patients with TLE specific to the lateralization of seizure focus, and describe compensatory mechanisms especially in the CH of patients with RTLE. We propose that graph theoretical analysis of metabolic connectivity using [(18)F]FDG-PET offers an important additional modality to explore brain networks.

  10. First Degree Pacemaker Exit Block

    Directory of Open Access Journals (Sweden)

    Johnson Francis

    2016-10-01

    Full Text Available Usually atrial and ventricular depolarizations follow soon after the pacemaker stimulus (spike on the ECG. But there can be an exit block due to fibrosis at the electrode - tissue interface at the lead tip. This can increase the delay between the spike and atrial or ventricular depolarization.

  11. A forced desynchrony study of circadian pacemaker characteristics in seasonal affective disorder

    NARCIS (Netherlands)

    Koorengevel, Kathelijne M.; Beersma, Domien G.M.; den Boer, Johan; Hoofdakker, Rutger H. van den

    2002-01-01

    The circadian pacemaker is an endogenous clock that regulates oscillations in most physiological and psychological processes with a near 24-h period. In many species, this pacemaker triggers seasonal changes in behavior. The seasonality of symptoms and the efficacy of light therapy suggest

  12. [Management of surgery patients with implanted cardiac pacemakers].

    Science.gov (United States)

    Ugljen, R; Dadić, D; Ferek-Petrić, B; Jelić, I; Letica, D; Anić, D; Husar, J

    1995-01-01

    Patients having cardiac pacemaker implanted may be subjected to various general surgery procedures. Application of electrosurgery for the purpose of resection and coagulation, provides a high frequency electric field which produces electric voltage on the electrodes of the pacing system. This voltage may be detected within the pacing system, and various arrhythmias can be provoked in correlation with underlying rhythm and mode of pacing. Preoperative patient control and proper pacemaker programming can prevent the pacing malfunctions due to the electrosurgery application. Appropriate positioning of the neutral electrode in relation to the pacing system avoids the electric fields intersection and decreases their interference.

  13. Echocardiographic findings predict in-hospital and 1-year mortality in left-sided native valve Staphylococcus aureus endocarditis

    DEFF Research Database (Denmark)

    Lauridsen, Trine K.; Park, Lawrence; Tong, Steven Y C

    2015-01-01

    BACKGROUND: Staphylococcus aureus left-sided native valve infective endocarditis (LNVIE) has higher complication and mortality rates compared with endocarditis from other pathogens. Whether echocardiographic variables can predict prognosis in S aureus LNVIE is unknown. METHODS AND RESULTS......: Consecutive patients with LNVIE, enrolled between January 2000 and September 2006, in the International Collaboration on Endocarditis were identified. Subjects without S aureus IE were matched to those with S aureus IE by the propensity of having S aureus. Survival differences were determined using log...

  14. Conduction Abnormalities and Permanent Pacemaker Implantation After Transcatheter Aortic Valve Replacement Using the Repositionable LOTUS Device: The United Kingdom Experience.

    Science.gov (United States)

    Rampat, Rajiv; Khawaja, M Zeeshan; Hilling-Smith, Roland; Byrne, Jonathan; MacCarthy, Philip; Blackman, Daniel J; Krishnamurthy, Arvindra; Gunarathne, Ashan; Kovac, Jan; Banning, Adrian; Kharbanda, Raj; Firoozi, Sami; Brecker, Stephen; Redwood, Simon; Bapat, Vinayak; Mullen, Michael; Aggarwal, Suneil; Manoharan, Ganesh; Spence, Mark S; Khogali, Saib; Dooley, Maureen; Cockburn, James; de Belder, Adam; Trivedi, Uday; Hildick-Smith, David

    2017-06-26

    The authors report the incidence of pacemaker implantation up to hospital discharge and the factors influencing pacing rate following implantation of the LOTUS bioprosthesis (Boston Scientific, Natick, Massachusetts) in the United Kingdom. Transcatheter aortic valve replacement (TAVR) is associated with a significant need for permanent pacemaker implantation. Pacing rates vary according to the device used. The REPRISE II (Repositionable Percutaneous Replacement of Stenotic Aortic Valve Through Implantation of Lotus Valve System) trial reported a pacing rate of 29% at 30 days after implantation of the LOTUS device. Data were collected retrospectively on 228 patients who had the LOTUS device implanted between March 2013 and February 2015 across 10 centers in the United Kingdom. Twenty-seven patients (12%) had pacemakers implanted pre-procedure and were excluded from the analysis. Patients were aged 81.2 ± 7.7 years; 50.7% were male. The mean pre-procedural QRS duration was 101.7 ± 20.4 ms. More than one-half of the cohort (n = 111, 55%) developed new left bundle branch block (LBBB) following the procedure. Permanent pacemakers were implanted in 64 patients (32%) with a median time to insertion of 3.0 ± 3.4 days. Chief indications for pacing were atrioventricular (AV) block (n = 46, 72%), or LBBB with 1st degree AV block (n = 11, 17%). Amongst those who received a pacemaker following TAVR the pre-procedural electrocardiogram findings included: No conduction disturbance (n = 41, 64%); 1st degree AV block (n = 10, 16%); right bundle branch block (n = 6, 9%) and LBBB (n = 5, 8%). LBBB (but not permanent pacemaker) occurred more frequently in patients who had balloon aortic valvuloplasty before TAVR (odds ratio [OR]: 1.25; p = 0.03). Pre-procedural conduction abnormality (composite of 1st degree AV block, hemiblock, right bundle branch block, LBBB) was independently associated with the need for permanent pacemaker (OR: 2.54; p = 0.048). The absence of

  15. Radiotherapy for breast cancer and pacemaker; Radiotherapie pour un cancer du sein et stimulateur cardiaque

    Energy Technology Data Exchange (ETDEWEB)

    Menard, J.; Campana, F.; Bollet, M.A.; Dendale, R.; Fournier-Bidoz, N.; Marchand, V.; Mazal, A.; Fourquet, A.; Kirova, Y.M. [Oncologie-radiotherapie, institut Curie, 26, rue d' Ulm, 75005 Paris (France); Kirov, K.M.; Esteve, M. [Departement d' anesthesie-reanimation-douleur, institut Curie, 75005 Paris (France)

    2011-06-15

    Purpose. - Patients with permanent cardiac pacemakers occasionally require radiotherapy. Therapeutic Irradiation may cause pacemakers to malfunction due to the effects of ionizing radiation or electromagnetic interference. We studied the breast cancer patients who needed breast and/or chest wall and lymph node irradiation to assess the feasibility and tolerance in this population of patients. Patients and methods. - From November 2008 to December 2009, more than 900 patients received radiotherapy for their breast cancer in our department using megavoltage linear accelerator (X 4-6 MV and electrons). Among them, seven patients were with permanent pacemaker. All patients have been treated to the breast and chest wall and/or lymph nodes. Total dose to breast and/or chest wall was 50 Gy/25 fractions and 46 Gy/23 fractions to lymph nodes. Patients who underwent conserving surgery followed by breast irradiation were boosted when indicated to tumour bed with 16 Gy/8 fractions. All patients were monitored everyday in presence of radiation oncologist to follow the function of their pacemaker. All pacemakers were controlled before and after radiotherapy by the patients' cardiologist. Results. - Seven patients were referred in our department for postoperative breast cancer radiotherapy. Among them, only one patient was declined for radiotherapy and underwent mastectomy without radiotherapy. In four cases the pacemaker was repositioned before the beginning of radiotherapy. Six patients, aged between 48 and 84 years underwent irradiation for their breast cancer. Four patients were treated with conserving surgery followed by breast radiotherapy and two with mastectomy followed by chest wall and internal mammary chain, supra- and infra-clavicular lymph node irradiation. The dose to the pacemaker generator was kept below 2 Gy. There was no pacemaker dysfunction observed during the radiotherapy. Conclusion. - The multidisciplinary work with position change of the pacemaker

  16. Pacemaker dependency after transcatheter aortic valve implantation with the self-expanding Medtronic CoreValve System.

    Science.gov (United States)

    van der Boon, Robert M A; Van Mieghem, Nicolas M; Theuns, Dominic A; Nuis, Rutger-Jan; Nauta, Sjoerd T; Serruys, Patrick W; Jordaens, Luc; van Domburg, Ron T; de Jaegere, Peter P T

    2013-09-30

    To determine pacemaker (PM) dependency at follow-up visit in patients who underwent new permanent pacemaker implantation (PPI) following transcatheter aortic valve implantation (TAVI). Single center prospective observational study including 167 patients without previous PM implantation who underwent TAVI with the self-expanding Medtronic CoreValve System (MCS) between November 2005 and February 2011. PM dependency was defined by the presence of a high degree atrioventricular block (HDAVB; second [AV2] and third degree [AV3B]), or a slow (atrial fibrilation with slow response (n=1, 2.8%) and left bundle branch block (n=1, 2.8%). Long term follow-up was complete for all patients and ranged from 1 to 40 months (median (IQR): 11.5 (5.0-18.0 months). Of those patients with a HDAVB, 16 out of the 30 patients (53.3%) were PM independent at follow-up visit (complete or partial resolution of the AV conduction abnormality). Overall, 20 out of the 36 patients (55.6%) who received a new PM following TAVI were PM independent at follow-up. Partial and even complete resolution of peri-operative AV conduction abnormalities after MCS valve implantation occurred in more than half of the patients. Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

  17. Pacemaker neuron and network oscillations depend on a neuromodulator-regulated linear current

    Directory of Open Access Journals (Sweden)

    Shunbing Zhao

    2010-05-01

    Full Text Available Linear leak currents have been implicated in the regulation of neuronal excitability, generation of neuronal and network oscillations, and network state transitions. Yet, few studies have directly tested the dependence of network oscillations on leak currents or explored the role of leak currents on network activity. In the oscillatory pyloric network of decapod crustaceans neuromodulatory inputs are necessary for pacemaker activity. A large subset of neuromodulators is known to activate a single voltage-gated inward current IMI, which has been shown to regulate the rhythmic activity of the network and its pacemaker neurons. Using the dynamic clamp technique, we show that the crucial component of IMI for the generation of oscillatory activity is only a close-to-linear portion of the current-voltage relationship. The nature of this conductance is such that the presence or the absence of neuromodulators effectively regulates the amount of leak current and the input resistance in the pacemaker neurons. When deprived of neuromodulatory inputs, pyloric oscillations are disrupted; yet, a linear reduction of the total conductance in a single neuron within the pacemaker group recovers not only the pacemaker activity in that neuron, but also leads to a recovery of oscillations in the entire pyloric network. The recovered activity produces proper frequency and phasing that is similar to that induced by neuromodulators. These results show that the passive properties of pacemaker neurons can significantly affect their capacity to generate and regulate the oscillatory activity of an entire network, and that this feature is exploited by neuromodulatory inputs.

  18. Left neglect dyslexia: Perseveration and reading error types.

    Science.gov (United States)

    Ronchi, Roberta; Algeri, Lorella; Chiapella, Laura; Gallucci, Marcello; Spada, Maria Simonetta; Vallar, Giuseppe

    2016-08-01

    Right-brain-damaged patients may show a reading disorder termed neglect dyslexia. Patients with left neglect dyslexia omit letters on the left-hand-side (the beginning, when reading left-to-right) part of the letter string, substitute them with other letters, and add letters to the left of the string. The aim of this study was to investigate the pattern of association, if any, between error types in patients with left neglect dyslexia and recurrent perseveration (a productive visuo-motor deficit characterized by addition of marks) in target cancellation. Specifically, we aimed at assessing whether different productive symptoms (relative to the reading and the visuo-motor domains) could be associated in patients with left spatial neglect. Fifty-four right-brain-damaged patients took part in the study: 50 out of the 54 patients showed left spatial neglect, with 27 of them also exhibiting left neglect dyslexia. Neglect dyslexic patients who showed perseveration produced mainly substitution neglect errors in reading. Conversely, omissions were the prevailing reading error pattern in neglect dyslexic patients without perseveration. Addition reading errors were much infrequent. Different functional pathological mechanisms may underlie omission and substitution reading errors committed by right-brain-damaged patients with left neglect dyslexia. One such mechanism, involving the defective stopping of inappropriate responses, may contribute to both recurrent perseveration in target cancellation, and substitution errors in reading. Productive pathological phenomena, together with deficits of spatial attention to events taking place on the left-hand-side of space, shape the manifestations of neglect dyslexia, and, more generally, of spatial neglect. Copyright © 2016 Elsevier Ltd. All rights reserved.

  19. Evaluation of cumulative effects of MR imaging on pacemaker systems at 1.5 Tesla.

    Science.gov (United States)

    Naehle, Claas P; Zeijlemaker, Volkert; Thomas, Daniel; Meyer, Carsten; Strach, Katharina; Fimmers, Rolf; Schild, Hans; Sommer, Torsten

    2009-12-01

    The purpose of this study was to evaluate possible cumulative effects of repeated magnetic resonance imaging (MRI) examinations on pacemaker systems in patients with cardiac pacemakers. The records of pacemaker patients who underwent repetitive MRI examinations in our institution were reviewed to identify patients who underwent two or more MRI examinations at 1.5T of any anatomical region. Using these criteria, a total of 47 patients who underwent a total 171 MRI examinations were identified and included in this study. Institutional Review Board approval for all pacemaker investigations was obtained. Written informed consent was obtained from all patients. Pacemakers were interrogated immediately before and after MR imaging, and after 3 months, including measurement of pacing capture threshold (PCT), lead impedance (LI), and battery voltage (BV). PCT, LI, and BV were analyzed for changes dependant on the number of MRI exams performed. Mean changes over time and changes between first and last pacemaker interrogation of PCT, LI, and BV were calculated. A statistically significant (P < 0.05), but clinically irrelevant trend for decrease in PCT and BV was found. No significant or clinically relevant changes in LI were observed. In this first study, no clinically relevant, cumulative changes in PCT, LI, or BV could be detected in PM patients who underwent two or more MRI examinations. However, a careful benefit/risk evaluation, among other MRI- and pacemaker-related safety precautions, remains mandatory, as clinically relevant alterations to the PM system cannot be excluded by all means.

  20. SU-E-I-42: Normalized Embryo/fetus Doses for Fluoroscopically Guided Pacemaker Implantation Procedures Calculated Using a Monte Carlo Technique

    Energy Technology Data Exchange (ETDEWEB)

    Damilakis, J; Stratakis, J; Solomou, G [University of Crete, Heraklion (Greece)

    2014-06-01

    Purpose: It is well known that pacemaker implantation is sometimes needed in pregnant patients with symptomatic bradycardia. To our knowledge, there is no reported experience regarding radiation doses to the unborn child resulting from fluoroscopy during pacemaker implantation. The purpose of the current study was to develop a method for estimating embryo/fetus dose from fluoroscopically guided pacemaker implantation procedures performed on pregnant patients during all trimesters of gestation. Methods: The Monte Carlo N-Particle (MCNP) radiation transport code was employed in this study. Three mathematical anthropomorphic phantoms representing the average pregnant patient at the first, second and third trimesters of gestation were generated using Bodybuilder software (White Rock science, White Rock, NM). The normalized embryo/fetus dose from the posteroanterior (PA), the 30° left-anterior oblique (LAO) and the 30° right-anterior oblique (RAO) projections were calculated for a wide range of kVp (50–120 kVp) and total filtration values (2.5–9.0 mm Al). Results: The results consist of radiation doses normalized to a) entrance skin dose (ESD) and b) dose area product (DAP) so that the dose to the unborn child from any fluoroscopic technique and x-ray device used can be calculated. ESD normalized doses ranged from 0.008 (PA, first trimester) to 2.519 μGy/mGy (RAO, third trimester). DAP normalized doses ranged from 0.051 (PA, first trimester) to 12.852 μGy/Gycm2 (RAO, third trimester). Conclusion: Embryo/fetus doses from fluoroscopically guided pacemaker implantation procedures performed on pregnant patients during all stages of gestation can be estimated using the method developed in this study. This study was supported by the Greek Ministry of Education and Religious Affairs, General Secretariat for Research and Technology, Operational Program ‘Education and Lifelong Learning’, ARISTIA (Research project: CONCERT)

  1. SU-E-I-42: Normalized Embryo/fetus Doses for Fluoroscopically Guided Pacemaker Implantation Procedures Calculated Using a Monte Carlo Technique

    International Nuclear Information System (INIS)

    Damilakis, J; Stratakis, J; Solomou, G

    2014-01-01

    Purpose: It is well known that pacemaker implantation is sometimes needed in pregnant patients with symptomatic bradycardia. To our knowledge, there is no reported experience regarding radiation doses to the unborn child resulting from fluoroscopy during pacemaker implantation. The purpose of the current study was to develop a method for estimating embryo/fetus dose from fluoroscopically guided pacemaker implantation procedures performed on pregnant patients during all trimesters of gestation. Methods: The Monte Carlo N-Particle (MCNP) radiation transport code was employed in this study. Three mathematical anthropomorphic phantoms representing the average pregnant patient at the first, second and third trimesters of gestation were generated using Bodybuilder software (White Rock science, White Rock, NM). The normalized embryo/fetus dose from the posteroanterior (PA), the 30° left-anterior oblique (LAO) and the 30° right-anterior oblique (RAO) projections were calculated for a wide range of kVp (50–120 kVp) and total filtration values (2.5–9.0 mm Al). Results: The results consist of radiation doses normalized to a) entrance skin dose (ESD) and b) dose area product (DAP) so that the dose to the unborn child from any fluoroscopic technique and x-ray device used can be calculated. ESD normalized doses ranged from 0.008 (PA, first trimester) to 2.519 μGy/mGy (RAO, third trimester). DAP normalized doses ranged from 0.051 (PA, first trimester) to 12.852 μGy/Gycm2 (RAO, third trimester). Conclusion: Embryo/fetus doses from fluoroscopically guided pacemaker implantation procedures performed on pregnant patients during all stages of gestation can be estimated using the method developed in this study. This study was supported by the Greek Ministry of Education and Religious Affairs, General Secretariat for Research and Technology, Operational Program ‘Education and Lifelong Learning’, ARISTIA (Research project: CONCERT)

  2. 76 FR 53851 - Effective Date of Requirement for Premarket Approval for Cardiovascular Permanent Pacemaker...

    Science.gov (United States)

    2011-08-30

    ... Pacemaker Electrode; Correction AGENCY: Food and Drug Administration, HHS. ACTION: Proposed rule; correction... preamendments device: Cardiovascular permanent pacemaker electrode. The document was published with an incorrect...

  3. Cardiac resynchronisation therapy in the presence of left-to-right intracardiac shunting: more good than harm?

    Science.gov (United States)

    Kyu, Kyu; Seow, Swee Chong; Wong, Raymond; Kojodjojo, Pipin

    2016-03-17

    An elderly Chinese man with moderately impaired left ventricular function, left bundle branch block and ST-elevation myocardial infarction complicated by ventricular septal rupture had class IV heart failure symptoms refractory to medical and surgical interventions. As a treatment of last resort, a cardiac resynchronisation therapy (CRT) pacemaker was implanted apprehensively, as preoperative concerns were raised whether CRT could exacerbate left-to-right shunting, hence negating the potential benefits of CRT. Introduction of CRT significantly improved the patient's haemodynamic status and symptoms, allowing for successful discharge home. To the best of our knowledge, this is the first report of a patient with severely symptomatic acute heart failure, widened QRS and active left-to-right intracardiac shunting, treated successfully with CRT. 2016 BMJ Publishing Group Ltd.

  4. Minimally Invasive Epicardial Pacemaker Implantation in Neonates with Congenital Heart Block.

    Science.gov (United States)

    Costa, Roberto; Silva, Katia Regina da; Martinelli Filho, Martino; Carrillo, Roger

    2017-10-01

    Few studies have characterized the surgical outcomes following epicardial pacemaker implantation in neonates with congenital complete atrioventricular block (CCAVB). This study sought to assess the long-term outcomes of a minimally invasive epicardial approach using a subxiphoid access for pacemaker implantation in neonates. Between July 2002 and February 2015, 16 consecutive neonates underwent epicardial pacemaker implantation due to CCAVB. Among these, 12 (75.0%) had congenital heart defects associated with CCAVB. The patients had a mean age of 4.7 ± 5.3 days and nine (56.3%) were female. Bipolar steroid-eluting epicardial leads were implanted in all patients through a minimally invasive subxiphoid approach and fixed on the diaphragmatic ventricular surface. The pulse generator was placed in an epigastric submuscular position. All procedures were successful, with no perioperative complications or early deaths. Mean operating time was 90.2 ± 16.8 minutes. None of the patients displayed pacing or sensing dysfunction, and all parameters remained stable throughout the follow-up period of 4.1 ± 3.9 years. Three children underwent pulse generator replacement due to normal battery depletion at 4.0, 7.2, and 9.0 years of age without the need of ventricular lead replacement. There were two deaths at 12 and 325 days after pacemaker implantation due to bleeding from thrombolytic use and progressive refractory heart failure, respectively. Epicardial pacemaker implantation through a subxiphoid approach in neonates with CCAVB is technically feasible and associated with excellent surgical outcomes and pacing lead longevity.

  5. SU-E-T-483: Treatment Planning Study of Volumetric Modulated Arc Therapy for Left-Sided Breast and Chestwall Cancers

    Energy Technology Data Exchange (ETDEWEB)

    Xu, H [Department of Radiation Oncology, Dalhousie University, Halifax, NS (Canada); Cape Breton Cancer Centre, Sydney, NS (Australia)

    2014-06-01

    Purpose: To perform the comparison of dose distributions and dosevolume- histograms generated by VMAT and conventional field-in-field technique for left-sided breast and chestwall cancers; to determine whether VMAT offers more dosimetric benefits than does the field-in-field technique. Methods: All VMAT and field-in-filed plans were produced in Eclipse(version 10). Five plans were generated for left-sided breast and leftsided chestwall with supraclavicular nodes, respectively. A clockwise arc (CW) and a counter-clockwise arc (CCW) were used with start and stop angles being 310o±10o and 140o±10o. Collimator angles were 30o for CW and 330o for CCW. The conformity index (CI) is the ratio of V95% over PTV. The homogeneity index (HI) is the ratio of the difference between D2% and D98% over the prescribed dose. The V5, as an indicator of low dose bath to organs-at-risk, was used for ipsilateral lung, heart, contralateral lung, and contralateral breast. The V20, as an indicator of radiation pneumonitis, was used for ipsilateral lung. Results: Breast/chestwall VMAT delivers much higher low dose bath to ipsilateral lung, contralateral lung and contralateral breast/chestwall for both intact breast and chestwall with nodes. V5 for heart is increased in VMAT plans. V20 for ipsilateral lung is lower in VMAT plans. PTV coverage is similar for both techniques. For one particular chestwall patient with supraclavicular and internal mammary nodes, VMAT offers superior dose coverage of PTVs with slightly more low-dose-wash to heart, contralateral lung and contralateral breast. Conclusion: This study indicates that there is generally no benefit using VMAT for left-sided intact breast, due to large low-dose-bath (5Gy) to normal tissues with insignificant improvement in PTV coverage. Dosimetric benefits will be seen in VMAT plans for some chestwall patients with large size, and/or internal mammary nodes, etc. Whether a chestwall patient is treated with VMAT should be carefully

  6. Anterior Myocardial Territory May Replace the Heart as Organ at Risk in Intensity-Modulated Radiotherapy for Left-Sided Breast Cancer

    International Nuclear Information System (INIS)

    Tan Wenyong; Liu Dong; Xue Chenbin; Xu Jiaozhen; Li Beihui; Chen Zhengwang; Hu Desheng; Wang Xionghong

    2012-01-01

    Purpose: We investigated whether the heart could be replaced by the anterior myocardial territory (AMT) as the organ at risk (OAR) in intensity-modulated radiotherapy (IMRT) of the breast for patients with left-sided breast cancer. Methods and Materials: Twenty-three patients with left-sided breast cancer who received postoperative radiation after breast-conserving surgery were studied. For each patient, we generated five IMRT plans including heart (H), left ventricle (LV), AMT, LV+AMT, and H+LV as the primary OARs, respectively, except both lungs and right breast, which corresponded to IMRT(H), IMRT(LV), IMRT(AMT), IMRT(LV+AMT), and IMRT(H+LV). For the planning target volumes and OARs, the parameters of dose–volume histograms were compared. Results: The homogeneity index, conformity index, and coverage index were not compromised significantly in IMRT(AMT), IMRT(LV) and IMRT(LV+ AMT), respectively, when compared with IMRT(H). The mean dose to the heart, LV, and AMT decreased 5.3–21.5% (p < 0.05), 19.9–29.5% (p < 0.05), and 13.3–24.5% (p < 0.05), respectively. Similarly, the low (e.g., V5%), middle (e.g., V20%), and high (e.g., V30%) dose-volume of the heart, LV, and AMT decreased with different levels. The mean dose and V10% of the right lung increased by 9.2% (p < 0.05) and 27.6% (p < 0.05), respectively, in IMRT(LV), and the mean dose and V5% of the right breast decreased significantly in IMRT(AMT) and IMRT(LV+AMT). IMRT(AMT) was the preferred plan and was then compared with IMRT(H+LV); the majority of dose–volume histogram parameters of OARs including the heart, LV, AMT, both lungs, and the right breast were not statistically different. However, the low dose-volume of LV increased and the middle dose-volume decreased significantly (p < 0.05) in IMRT(AMT). Also, those of the right lung (V10%, V15%) and right breast (V5%, V10%) decreased significantly (p < 0.05). Conclusions: The AMT may replace the heart as the OAR in left-sided breast IMRT after breast

  7. Gene- and cell-based bio-artificial pacemaker: what basic and translational lessons have we learned?

    Science.gov (United States)

    Li, R A

    2012-06-01

    Normal rhythms originate in the sino-atrial node, a specialized cardiac tissue consisting of only a few thousands of nodal pacemaker cells. Malfunction of pacemaker cells due to diseases or aging leads to rhythm generation disorders (for example, bradycardias and sick-sinus syndrome (SSS)), which often necessitate the implantation of electronic pacemakers. Although effective, electronic devices are associated with such shortcomings as limited battery life, permanent implantation of leads, lead dislodging, the lack of autonomic responses and so on. Here, various gene- and cell-based approaches, with a particular emphasis placed on the use of pluripotent stem cells and the hyperpolarization-activated cyclic nucleotide-gated-encoded pacemaker gene family, that have been pursued in the past decade to reconstruct bio-artificial pacemakers as alternatives will be discussed in relation to the basic biological insights and translational regenerative potential.

  8. Leadless pacemaker extraction from a single-center perspective.

    Science.gov (United States)

    González Villegas, Elkin; Al Razzo, Omar; Silvestre García, Jorge; Mesa García, José

    2018-02-01

    Leadless pacemaker can be considered as a technical revolution in cardiac pacing devices, with clear advantages over conventional pacemakers in overcoming all lead-related complications. However, the management of these devices once they reach the end of life (EOL) of the battery is still controversial. In the next years, there will be an increase in the need to define a clear strategy in the management of leadless PM once they reach their EOL. Safe extraction of these devices will define in a great manner this strategy METHODS: We performed the extraction of three functioning Nanostim leadless pacemaker prophylactically in two females and one male patients as part of the Nanostim battery depletion field action recommendation. All patients had a prior transesophageal 3D echocardiography to determine the device intracardiac mobility and the extent of possible endothelialization. For the extractions, we used the Nanostim Retrieval Catheter S1RSIN (St. Jude Medical, St. Paul, MN, USA), which is a proprietary catheter provided by the manufacturing company based on a lasso. Complete extraction of the devices was achieved in all patients using a relatively short fluoroscopic time (16, 19, and 12 minutes). The extraction of leadless pacemakers can be considered as a safe and feasible procedure using the tools provided by the manufacturer and designed for the extraction. However, a very low threshold must be maintained to avoid any risk to the patients. Our extraction time ranges are between 983 and 1,070 days, nevertheless it is necessary to gather more long-term data to assess the feasibility and safety of these procedures. © 2017 Wiley Periodicals, Inc.

  9. Transvenous permanent pacemaker implantation in dextrocardia: technique, challenges, outcome, and a brief review of literature.

    Science.gov (United States)

    Shenthar, Jayaprakash; Rai, Maneesh K; Walia, Rohit; Ghanta, Somasekhar; Sreekumar, Praveen; Reddy, Satish S

    2014-09-01

    Dextrocardia is a rare congenital anomaly. Pacemaker implantation in dextrocardia can be challenging because of the distorted anatomy and associated anomalies. The literature regarding implantation of pacemaker in dextrocardia is scarce. The study involved retrospective analysis of records of patients with dextrocardia who had undergone pacemaker implantation between January 2006 and July 2013 from a single centre. Six patients with dextrocardia (five males and one female) underwent permanent pacemaker implantation (PPI) between January 2006 and July 2013. Of them, three had situs solitus dextrocardia and three situs inversus dextrocardia. All three patients with situs solitus dextrocardia had associated corrected transposition of great arteries. The indication for pacemaker implantation was symptomatic complete atrioventricular (AV) block in four, high-grade AV block in one, and sinus node dysfunction in one patient. A favourable outcome was noted during a mean follow-up of 3.9 years (4 months to 7 years) with one patient needing a pulse generator replacement. Permanent pacemaker implantation in dextrocardia can be challenging because of the distorted anatomy. Use of a technique employing angiography to delineate chamber anatomy and relationship can assist the operator during such difficult PPIs. The medium- and long-term survival after a successful pacemaker implantation in dextrocardia is favourable. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

  10. ARTERIAL HYPERTENSION AND MEDICAL SUPPORT OF PATIENTS WITH PERMANENT PACEMAKERS

    Directory of Open Access Journals (Sweden)

    T. A. Derienko

    2016-06-01

    Full Text Available The review is devoted to clinical problems of arterial hypertension (AH in patients with implanted pacemakers (EKS and cardiac resynchronization therapy (CRT. Indications for pacemaker implantation and CRT are considered, especially the purpose and effectiveness of angiotensin-converting enzyme (ACE inhibitors, angiotensin II receptor antagonists (ARA, sartans, beta-blockers (BAB, diuretics, calcium channel blockers. We prove that the CRT and cardiac pacing and do not cancel, bur modify drug therapy of AH.

  11. Long-Term Mortality Effect of Early Pacemaker Implantation After Surgical Aortic Valve Replacement.

    Science.gov (United States)

    Greason, Kevin L; Lahr, Brian D; Stulak, John M; Cha, Yong-Mei; Rea, Robert F; Schaff, Hartzell V; Dearani, Joseph A

    2017-10-01

    The need for pacemaker implantation is a well-described complication of aortic valve replacement. Not so well described is the effect such an event has on long-term outcome. This study reviewed a 21-year experience at the Mayo Clinic (Rochester, Minnesota) with aortic valve replacement to understand the influence of early postoperative pacemaker implantation on long-term mortality rates more clearly. This study retrospectively reviewed the records of 5,842 patients without previous pacemaker implantation who underwent surgical aortic valve replacement from January 1993 through June 2014. The median age of these patients was 73 years (range, 65 to 79 years), the median ejection fraction was 62% (range, 53% to 68%), 3,853 patients were male (66%), and coronary artery bypass graft operation was performed in 2,553 (44%) of the patients studied. Early pacemaker implantation occurred in 146 patients (2.5%) within 30 days of surgical aortic valve replacement. The median follow-up of patients was 11.1 years (range, 5.8 to 16.5 years), and all-cause mortality rates were 2.4% at 30 days, 6.4% at 1 year, 23.1% at 5 years, 48.3% at 10 years, and 67.9% at 15 years postoperatively. Early pacemaker implantation was associated with an increased risk of death after multivariable adjustment for baseline patients' characteristics (hazard ratio, 1.49; 95% confidence interval, 1.20, 1.84; p pacemaker implantation as a complication of surgical aortic valve replacement is associated with an increased risk of long-term death. Valve replacement-related pacemaker implantation rates should be important considerations with respect to new valve replacement paradigms, especially in younger and lower-risk patients. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  12. Ectopic jejunal pacemakers and gastric emptying after Roux gastrectomy: Effect of intestinal pacing

    International Nuclear Information System (INIS)

    Karlstrom, L.; Kelly, K.A.

    1989-01-01

    The aims of this study were to determine whether ectopic pacemakers are present after meals in the Roux limbs of dogs after vagotomy and Roux gastrectomy, whether these pacemakers slow gastric emptying of liquids or solids, and whether abolishing the pacemakers with electric pacing might speed any slow emptying that occurs. In six dogs that underwent vagotomy and Roux gastrectomy and in four dogs that underwent vagotomy and Billroth gastrectomy (controls), myoelectric activity of the Roux limb or duodenum was measured during gastric emptying of a 500 kcal mixed meal of 99mTc-labeled cooked egg and 111In-labeled milk. Roux dogs were tested with and without pacing of the Roux limb. Roux dogs showed ectopic pacemaker in the Roux limb that drove the pacesetter potentials of the limb in a reverse, or orad, direction during 57% of the postprandial recordings. Billroth dogs had no ectopic pacemakers (p less than 0.05). Liquids emptied more slowly in Roux dogs (half-life (t1/2) = 121 +/- 15 minutes) than in Billroth dogs (t1/2 = 43 +/- 9 minutes; p less than 0.05), but solids emptied similarly in both groups of dogs (t1/2 approximately 8 hours). Pacing the Roux limb abolished the ectopic pacemakers, restored the slow emptying of liquids to the more rapid rate found in the Billroth dogs (t1/2: paced Roux, 72 +/- 15 minutes; Billroth, 43 +/- 9 minutes; p greater than 0.05) and did not change emptying of solids. The conclusion was that ectopic pacemakers present in the Roux limb after vagotomy and Roux gastrectomy drove the limb in a reverse direction and slowed emptying of liquids after the operation. The defect was corrected by pacing the Roux limb in a forward direction

  13. Ectopic jejunal pacemakers and gastric emptying after Roux gastrectomy: Effect of intestinal pacing

    Energy Technology Data Exchange (ETDEWEB)

    Karlstrom, L.; Kelly, K.A. (Mayo Clinic, Rochester, MN (USA))

    1989-11-01

    The aims of this study were to determine whether ectopic pacemakers are present after meals in the Roux limbs of dogs after vagotomy and Roux gastrectomy, whether these pacemakers slow gastric emptying of liquids or solids, and whether abolishing the pacemakers with electric pacing might speed any slow emptying that occurs. In six dogs that underwent vagotomy and Roux gastrectomy and in four dogs that underwent vagotomy and Billroth gastrectomy (controls), myoelectric activity of the Roux limb or duodenum was measured during gastric emptying of a 500 kcal mixed meal of 99mTc-labeled cooked egg and 111In-labeled milk. Roux dogs were tested with and without pacing of the Roux limb. Roux dogs showed ectopic pacemaker in the Roux limb that drove the pacesetter potentials of the limb in a reverse, or orad, direction during 57% of the postprandial recordings. Billroth dogs had no ectopic pacemakers (p less than 0.05). Liquids emptied more slowly in Roux dogs (half-life (t1/2) = 121 +/- 15 minutes) than in Billroth dogs (t1/2 = 43 +/- 9 minutes; p less than 0.05), but solids emptied similarly in both groups of dogs (t1/2 approximately 8 hours). Pacing the Roux limb abolished the ectopic pacemakers, restored the slow emptying of liquids to the more rapid rate found in the Billroth dogs (t1/2: paced Roux, 72 +/- 15 minutes; Billroth, 43 +/- 9 minutes; p greater than 0.05) and did not change emptying of solids. The conclusion was that ectopic pacemakers present in the Roux limb after vagotomy and Roux gastrectomy drove the limb in a reverse direction and slowed emptying of liquids after the operation. The defect was corrected by pacing the Roux limb in a forward direction.

  14. Truck side guard specifications : recommended standard

    Science.gov (United States)

    2016-09-01

    This document is intended to be used by (1) public or private medium/heavy-duty truck fleets considering adding side guards; (2) jurisdictions or customers that require side guards through policy or procurement; (3) manufacturers of side guards; and ...

  15. Anaesthetic management of emergency pacemaker implantation in a case of neonatal lupus erythematosus with complete congenital heart block & severe respiratory distress

    Directory of Open Access Journals (Sweden)

    Usha Kiran

    2007-01-01

    Full Text Available An 8-week old 3-kilogram male baby was brought to this tertiary care hospital with respiratory distress, marked tracheal tug, poor feeding and a heart rate of 46/minute. The child had been referred from a peripheral hospital as a case of neonatal lupus with complete congenital heart block. The mother was seropositive for systemic lupus erythematosus with a history of two abortions. Evaluation on admission revealed a heart rate between 40-60/ minute, respiratory rate 40-50/ minute, inspiratory stridor, bilateral crepitations, chest retrac-tion and a marked tracheal tug that improved with prone positioning. Electrocardiography and echocardiography confirmed complete congenital heart block with cardiomegaly and mild left ventricular dysfunction. Keeping in view the impending congestive heart failure, possible early cardiomyopathy and the bad obstetric history ur-gent pacemaker implantation was planned to allow early recovery of the child. The anaesthetic risk was high due to the heart block, ventricular dysfunction, laryngomalacia, severe tracheal tug and anticipated difficult weaning from controlled ventilation. General anaesthesia was administered with endotracheal tube and con-trolled ventilation using ketamine, rocuronium and sufentanil. For patient safety invasive monitoring was pro-vided and external pacing was kept standby. Epicardial pacemaker leads were implanted onto the left ventricu-lar wall through a left anterior 6th intercostal space thoracotomy. The child was electively ventilated for two post operative days. The tracheal tug and secretions gradually subsided over 2 weeks with oxygen, antibiotics, steroids, bronchodilators and physiotherapy. At the time of discharge from hospital 2 weeks after the implant the child was feeding well, tracheal tug was minimal and the lungs were clear.

  16. Pacemaker-driven stochastic resonance on diffusive and complex networks of bistable oscillators

    Energy Technology Data Exchange (ETDEWEB)

    Perc, Matjaz; Gosak, Marko [Department of Physics, Faculty of Natural Sciences and Mathematics, University of Maribor, Koroska cesta 160, SI-2000 Maribor (Slovenia)], E-mail: matjaz.perc@uni-mb.si

    2008-05-15

    We study the phenomenon of stochastic resonance on diffusive, small-world and scale-free networks consisting of bistable overdamped oscillators. Important thereby is the fact that the external subthreshold periodic forcing is introduced only to a single oscillator of the network. Hence, the forcing acts as a pacemaker trying to impose its rhythm on the whole network through the unit to which it is introduced. Without the addition of additive spatiotemporal noise, however, the whole network, including the unit that is directly exposed to the pacemaker, remains trapped forever in one of the two stable steady states of the local dynamics. We show that the correlation between the frequency of subthreshold pacemaker activity and the response of the network is resonantly dependent on the intensity of additive noise. The reported pacemaker-driven stochastic resonance depends most significantly on the coupling strength and the underlying network structure. Namely, the outreach of the pacemaker obeys the classic diffusion law in the case of nearest-neighbor interactions, thus being proportional to the square root of the coupling strength, whereas it becomes superdiffusive by an appropriate small-world or scale-free topology of the interaction network. In particular, the scale-free topology is identified as being optimal for the dissemination of localized rhythmic activity across the whole network. Also, we show that the ratio between the clustering coefficient and the characteristic path length is the crucial quantity defining the ability of a small-world network to facilitate the outreach of the pacemaker-emitted subthreshold rhythm. We additionally confirm these findings by using the FitzHugh-Nagumo excitable system as an alternative to the bistable overdamped oscillator.

  17. Pacemaker-driven stochastic resonance on diffusive and complex networks of bistable oscillators

    International Nuclear Information System (INIS)

    Perc, Matjaz; Gosak, Marko

    2008-01-01

    We study the phenomenon of stochastic resonance on diffusive, small-world and scale-free networks consisting of bistable overdamped oscillators. Important thereby is the fact that the external subthreshold periodic forcing is introduced only to a single oscillator of the network. Hence, the forcing acts as a pacemaker trying to impose its rhythm on the whole network through the unit to which it is introduced. Without the addition of additive spatiotemporal noise, however, the whole network, including the unit that is directly exposed to the pacemaker, remains trapped forever in one of the two stable steady states of the local dynamics. We show that the correlation between the frequency of subthreshold pacemaker activity and the response of the network is resonantly dependent on the intensity of additive noise. The reported pacemaker-driven stochastic resonance depends most significantly on the coupling strength and the underlying network structure. Namely, the outreach of the pacemaker obeys the classic diffusion law in the case of nearest-neighbor interactions, thus being proportional to the square root of the coupling strength, whereas it becomes superdiffusive by an appropriate small-world or scale-free topology of the interaction network. In particular, the scale-free topology is identified as being optimal for the dissemination of localized rhythmic activity across the whole network. Also, we show that the ratio between the clustering coefficient and the characteristic path length is the crucial quantity defining the ability of a small-world network to facilitate the outreach of the pacemaker-emitted subthreshold rhythm. We additionally confirm these findings by using the FitzHugh-Nagumo excitable system as an alternative to the bistable overdamped oscillator

  18. The Effects Of Short Tıme Rıght Ventrıcular Pacıng On Left Atrıal Mechanıcal Functıons

    Directory of Open Access Journals (Sweden)

    Oyku Gulmez

    2009-05-01

    Full Text Available Objectives: Left atrium (LA plays an important role in left ventricular filling. It is well known that right ventricular apical pacing has unfavorable effects on ventricular systolic and diastolic performance. The aim of this study is to evaluate the LA mechanical functions with 2D echocardiography in patients with a permanent pacemaker after short time ventricular pacing.Design: Echocardiographic examination was performed in 38 patients (mean age 63.0±10.9, 18 female with dual chamber pacemakers or defibrillators (90% ventricular pacing at 70 beats per minute in DDD mode with an optimal AV interval. Left atrial volumes (LAV including at the time of mitral valve opening (Vmax, at closure (Vmin, and at the onset of atrial systole (Volp were measured. The passive emptying, conduit, active emptying and total emptying volume, stroke volumes were also calculated. Results: No significant differences were noted at baseline and after pacing for absolute Vmax, Volp, passive emptying, conduit, active emptying, total emptying volumes as well as the volumes indexed to body surface area (p >0.05. Conclusion: Short - time RV pacing seems to have no acute effects on left atrial mechanical functions.

  19. Frequency of pacemaker malfunction associated with monopolar electrosurgery during pulse generator replacement or upgrade surgery.

    Science.gov (United States)

    Lin, Yun; Melby, Daniel P; Krishnan, Balaji; Adabag, Selcuk; Tholakanahalli, Venkatakrishna; Li, Jian-Ming

    2017-08-01

    The aim of this study is to investigate the frequency of electrosurgery-related pacemaker malfunction. A retrospective study was conducted to investigate electrosurgery-related pacemaker malfunction in consecutive patients undergoing pulse generator (PG) replacement or upgrade from two large hospitals in Minneapolis, MN between January 2011 and January 2014. The occurrence of this pacemaker malfunction was then studied by using MAUDE database for all four major device vendors. A total of 1398 consecutive patients from 2 large tertiary referral centers in Minneapolis, MN undergoing PG replacement or upgrade surgery were retrospectively studied. Four patients (0.3% of all patients), all with pacemakers from St Jude Medical (2.8%, 4 of 142) had output failure or inappropriately low pacing rate below 30 bpm during electrosurgery, despite being programmed in an asynchronous mode. During the same period, 1174 cases of pacemaker malfunctions were reported on the same models in MAUDE database, 37 of which (3.2%) were electrosurgery-related. Twenty-four cases (65%) had output failure or inappropriate low pacing rate. The distribution of adverse events was loss of pacing (59.5%), reversion to backup pacing (32.4%), inappropriate low pacing rate (5.4%), and ventricular fibrillation (2.7%). The majority of these (78.5%) occurred during PG replacement at ERI or upgrade surgery. No electrosurgery-related malfunction was found in MAUDE database on 862 pacemaker malfunction cases during the same period from other vendors. Electrosurgery during PG replacement or upgrade surgery can trigger output failure or inappropriate low pacing rate in certain models of modern pacemakers. Cautions should be taken for pacemaker-dependent patients.

  20. SU-E-T-532: Left-Sided Breast Cancer Irradiation Using Volumatric Modulated Arc Therapy: An Evaluation of Multiple Commercial Systems

    International Nuclear Information System (INIS)

    Liu, R; Liu, T; Qi, S

    2015-01-01

    Purposes: There has been growing interest in treating breast cancer using VMAT technique. Our goal is to compare the dosimetry and treatment delivery parameters for the left-sided breast cancer treatment using various VMAT platforms from commercially available planning systems. Methods: Five consecutive left-sided breast cancer patients initially treated with conventional 3D-conformal radiotherapy (3DCRT) were selected. Four VMAT plans using most popular treatment planning systems, including Eclipse (Version 11, Varian), Pinnacle (Version 9.8, Philips), Monaco (Version 2.03, Elekta) and helical Tomotherapy (V4.0, Accuray). The same structure set and same planning goals were used for all VMAT plans. The dosimetric parameters including target coverage and minimum/maximum/mean, dose-volume endpoints for the selected normal structures: the heart, ipsilateral-/contralateral lung and breast, were evaluated. Other dosimetric indices including heterogeneity index (HI) were evaluated. The treatment delivery parameters, such as monitor unit (MUs) and delivery time were also compared. Results: VMAT increases dose homogeneity to the treated volume and reduces the irradiated heart and left-lung volumes. Compared to the 3DCRT technique, all VMAT plans offer better heart and left-lung dose sparing; the mean heart doses were 4.5±1.6(Monaco), 1.2±0.4(Pinnacle), 1.3± (Eclipse) and 5.6±4.4(Tomo), the mean left-lung doses were 5.9±1.5(Monaco), 3.7±0.7(Pinnacle), 1.4± (Eclipse) and 5.2±1.6 (Tomo), while for the 3DCRT plan, the mean heart and left-Lung doses were 2.9±2.0, and 6.8±4.4 (Gy) respectively. The averaged contralateral-breast and lung mean doses were higher in VMAT plans than the 3DCRT plans but were not statistically significant. Among all the VMAT plans, the Pinnacle plans often yield the lowest right-lung/breast mean doses, and slightly better heterogeneity indices that are similar to Tomotherapy plans. Treatment delivery time of the VMAT plans (except helical

  1. A diagnostic dilemma: Left-sided appendicitis in a 10 year old boy with previously undiagnosed intestinal malrotation. A case report

    Directory of Open Access Journals (Sweden)

    Ashvini Shekhar

    2015-01-01

    Conclusion: Left sided acute appendicitis is a diagnostic dilemma, thus often leading to management delays. It is pertinent to remember that malrotation of the gut is more common than previously thought, and not just a disease of infancy. It is advisable to consider imaging studies while balancing the risk-benefit-ratio of radiation exposure, especially in paediatric cases to cinch the diagnosis.

  2. Echocardiography in patients with complications related to pacemakers and cardiac defibrillators.

    Science.gov (United States)

    Almomani, Ahmed; Siddiqui, Khadija; Ahmad, Masood

    2014-03-01

    The evolving indications and uses for implantable cardiac devices have led to a significant increase in the number of implanted devices each year. Implantation of endocardial leads for permanent pacemakers and cardiac defibrillators can cause many delayed complications. Complications may be mechanical and related to the interaction of the device leads with the valves and endomyocardium, e.g., perforation, infection, and thrombosis, or due to the electrical pacing of the myocardium and conduction abnormalities, e.g., left ventricular dyssynchrony. Tricuspid regurgitation, another delayed complication in these patients, may be secondary to both mechanical and pacing effects of the device leads. Echocardiography plays an important role in the diagnosis of these device-related complications. Both two-dimensional transthoracic echocardiography and transesophageal echocardiography provide useful diagnostic information. Real time three-dimensional echocardiography is a novel technique that can further enhance the detection of lead-related complications. © 2013, Wiley Periodicals, Inc.

  3. Segmental omental infarction in childhood: an unusual case of left-sided location with extension into the pelvis

    International Nuclear Information System (INIS)

    Foscolo, Sylvain; Mandry, Damien; Galloy, Marie-Agnes; Claudon, Michel; Champigneulles, Jacqueline; Miscault, Godefroy de

    2007-01-01

    Segmental omental infarction (OI) is a rare cause of abdominal pain in children. It generally occurs in the right lower and upper quadrants of the abdomen and only a few cases of other locations have been described in adults. We report a unique paediatric case of OI with an unusual left-sided location extending into the pelvis in a 6-year-old non-obese girl. The diagnosis was suspected on US and CT based on imaging patterns previously described in adults and children. Laparoscopic resection should be balanced with nonsurgical management. (orig.)

  4. Exercise thallium-201 myocardial imaging in left main coronary artery disease: sensitive but not specific

    International Nuclear Information System (INIS)

    Rehn, T.; Griffith, L.S.; Achuff, S.C.; Bailey, I.K.; Bulkley, B.H.; Burow, R.; Pitt, B.; Becker, L.C.

    1981-01-01

    To determine the usefulness of thallium-201 scintigraphy for identifying left main coronary artery disease, the results of scintigraphy at rest and during exercise were compared in 24 patients with 50 percent or greater narrowing of the left main coronary artery and 80 patients with 50 percent or greater narrowing of one or more of the major coronary arteries but without left main coronary involvement. By segmental analysis of the scintigrams, perfusion defects were assigned to the left anterior descending, left circumflex or right coronary artery, singly or in combination, and the pattern of simultaneous left anterior descending and circumflex arterial defects was used to identify left main coronary artery disease. Of the 24 patients with left main coronary artery disease, 22 (92 percent) had abnormal exercise scintigrams. Despite this high sensitivity, the pattern of perfusion defects was not specific; the ''left main pattern'' was found in 3 patients (13 percent) with left main coronary artery disease but also in 3 (33 percent) of 9 patients with combined left anterior descending and left circumflex arterial disease, 4 (19 percent) of 21 patients with three vessel disease and 3 (6 percent) of 50 patients with one or two vessel disease but excluding the group with left anterior descending plus left circumflex arterial disease. The pattern of perfusion defects in the patients with left main coronary artery disease was determined by the location and severity of narrowings in the coronary arteries downstream from the left main arterial lesion. Concomitant lesions in other arteries were found in all patients with left main coronary disease (one vessel in 1 patient, two vessels in 7 patients and three vessels in 16). For this reason, it is unlikely that even with improvements in radiopharmaceutical agents and imaging techniques, myocardial perfusion scintigraphy will be sufficiently specific for definitive identification of left main coronary artery disease

  5. Usefulness of Tricuspid Annular Diameter to Predict Late Right Sided Heart Failure in Patients With Left Ventricular Assist Device.

    Science.gov (United States)

    Nakanishi, Koki; Homma, Shunichi; Han, Jiho; Takayama, Hiroo; Colombo, Paolo C; Yuzefpolskaya, Melana; Garan, Arthur R; Farr, Maryjane A; Kurlansky, Paul; Di Tullio, Marco R; Naka, Yoshifumi; Takeda, Koji

    2018-07-01

    Although late-onset right-sided heart failure is recognized as a clinical problem in the treatment of patients with left ventricular assist devices (LVADs), the mechanism and predictors are unknown. Tricuspid valve (TV) deformation leads to the restriction of the leaflet motion and decreased coaptation, resulting in a functional tricuspid regurgitation that may act as a surrogate marker of late right-sided heart failure. This study aimed to investigate the association of preoperative TV deformation (annulus dilatation and leaflet tethering) with late right-sided heart failure development after continuous-flow LVAD implantation. The study cohort consisted of 274 patients who underwent 2-dimensional echocardiography before LVAD implantation. TV annulus diameter and tethering distance were measured in an apical 4-chamber view. Late right-sided heart failure was defined as right-sided heart failure requiring readmission and medical and/or surgical treatment after initial LVAD implantation. During a mean follow-up of 25.1 ± 19.0 months after LVAD implantation, late right-sided heart failure occurred in 33 patients (12.0%). Multivariate Cox proportional hazard analysis demonstrated that TV annulus diameter (hazard ratio 1.221 per 1 mm, p right-sided heart failure development, whereas leaflet tethering distance was not. The best cut-off value of the TV annular diameter was 41 mm (area under the curve 0.787). Kaplan-Meier analysis showed that patients with dilated TV annulus (TV annular diameter ≥41 mm) exhibited a significantly higher late right-sided heart failure occurrence than those without TV annular enlargement (log-rank p right-sided heart failure after LVAD implantation. Copyright © 2018 Elsevier Inc. All rights reserved.

  6. Social aspects of left-handedness

    Directory of Open Access Journals (Sweden)

    Belojević Goran

    2010-01-01

    Full Text Available Throughout human history left-handedness has been considered as sinful. It has been associated with the devil, weakness, female gender, unhealthiness, evil, something that has to be turned to a “good” - right side by force. Left-handedness is being more and more acceptable at rational level, but in everyday life it is still considered to be unusual if someone writes with the left hand. Lessening of the number of lefthanders is associated with ageing. There are about 13% lefthanders among people in twenties and less than 1% lefthanders among those in eighties. This finding may be explaned with more pronounced socio-cultural pressure on left-handed people in the past, compared to nowadays. On the other hand, this may also support the hypothesis about a reduced life span of lefthanded people. With cross-exercising of left-handedness, certain typical characteristics and behavioral patterns appear in these people. This was a sort of provoked behavior and an attack on the integrity of an emotional attitude toward oneself. Stuttering may also appear as a consequence of unsuccessful cross-exercising of left-handedness. The hypothesis about left-handedness as an advantage is supported with the reports about relatively more lefthanders in some specific groups such as: mathematicians, sculptors, architects, painters, musicians, actors, tennis players, as well as famous army commanders and rulers.

  7. Pacemaker replacement in nonagenarians: Procedural safety and long-term follow-up.

    Science.gov (United States)

    Loirat, Aurélie; Fénéon, Damien; Behaghel, Albin; Behar, Nathalie; Le Helloco, Alain; Mabo, Philippe; Daubert, Jean-Claude; Leclercq, Christophe; Martins, Raphaël P

    2015-01-01

    The rate of pacemaker implantation is rising. Given that the life expectancy of the population is projected to increase, a large number of elderly patients are likely to be implanted in the future. As pacemaker batteries can last for 8-10years, an increasing number of pacemaker recipients will require replacement of their devices when they become nonagenarians. To analyse the short- and long-term outcomes after device replacement in nonagenarians. Patients aged≥90years referred to a tertiary centre for pacemaker replacement from January 2004 to July 2014 were included retrospectively. Clinical follow-up data were obtained from clinical visits or telephone interviews with patients or their families. The primary clinical endpoint was total mortality. Secondary endpoints included early and delayed procedure-related complications and predictive risk factors for total mortality. Sixty-two patients were included (mean age 93.3±2.9years at time of pacemaker replacement). Mean procedure duration was 35.7±17.2minutes. Mean hospital stay was 2.2±1.1days. One patient died from a perioperative complication. Thirty-seven patients (59.7%) died during a median follow-up of 22.1months (interquartile range, 11.8-39.8months). Survival rates were 84.2% (95% confidence interval [CI] 71.8-91.5%) at 1year, 66.9% (95% CI 51.8-78.2%) at 2years and 22.7% (95% CI 10.6-37.7%) at 5years. Atrial fibrillation (hazard ratio 2.47, 95% CI 1.1-5.6) and non-physiological pacing (i.e. VVI pacing in patients in sinus rhythm) (hazard ratio 2.20, 95% CI 1.0-4.9) were predictors of mortality. Pacemaker replacement in nonagenarians is a safe and straightforward procedure. These data suggest that procedures can be performed securely in this old and frail population, with patients living for a median of 30months afterwards. Copyright © 2015 Elsevier Masson SAS. All rights reserved.

  8. Influence of ionizing radiation on the probability of failure of long-term pacemakers

    Energy Technology Data Exchange (ETDEWEB)

    Eger, H.; Hesse, W.; Otte, K.B.; Rauh, G.; Salewski, D.; Stopp, G. (Zentralklinik fuer Herz- und Lungenkrankheiten, Bad Berka (German Democratic Republic); VEB Transformatoren- und Roentgenwerk ' Hermann Matern' Dresden (German Democratic Republic); Martin-Luther-Universitaet Halle-Wittenberg, Halle (German Democratic Republic). Radiologische Klinik; Staatliches Amt fuer Atomsicherheit und Strahlenschutz, Berlin (German Democratic Republic))

    1982-11-01

    The functional disturbances observed during the exposure of cardiac pacemakers with a CMOS switching circle U 115 are described qualitatively for different types of radiation. Based on the maximum sensitivity of the U 115 within the X-ray range, the correlation for this range between dose of exposure and parameter changes is established. As is demonstrated such changes is established. As is demonstrated such changes are not clinically relevant as long as the dose of 5,16 x 10/sup -2/ C/kg is not exceeded. Proceeding from the accumulated dose to which the U 115 is exposed during various diagnostic radiographic procedures it is shown that such procedures do not result in functional disturbances of the pacemaker. Recommendations are given for radiotherapeutic measures as to pacemaker patients, which should be realized in close co-operation between the pacemaker centre concerned and an authorized hospital.

  9. Reciprocal Signaling between the Ectoderm and a Mesendodermal Left-Right Organizer Directs Left-Right Determination in the Sea Urchin Embryo

    Science.gov (United States)

    Bessodes, Nathalie; Haillot, Emmanuel; Duboc, Véronique; Röttinger, Eric; Lahaye, François; Lepage, Thierry

    2012-01-01

    During echinoderm development, expression of nodal on the right side plays a crucial role in positioning of the rudiment on the left side, but the mechanisms that restrict nodal expression to the right side are not known. Here we show that establishment of left-right asymmetry in the sea urchin embryo relies on reciprocal signaling between the ectoderm and a left-right organizer located in the endomesoderm. FGF/ERK and BMP2/4 signaling are required to initiate nodal expression in this organizer, while Delta/Notch signaling is required to suppress formation of this organizer on the left side of the archenteron. Furthermore, we report that the H+/K+-ATPase is critically required in the Notch signaling pathway upstream of the S3 cleavage of Notch. Our results identify several novel players and key early steps responsible for initiation, restriction, and propagation of left-right asymmetry during embryogenesis of a non-chordate deuterostome and uncover a functional link between the H+/K+-ATPase and the Notch signaling pathway. PMID:23271979

  10. Right Versus Left Colon Cancer Biology: Integrating the Consensus Molecular Subtypes.

    Science.gov (United States)

    Lee, Michael S; Menter, David G; Kopetz, Scott

    2017-03-01

    Although clinical management of colon cancer generally has not accounted for the primary tumor site, left-sided and right-sided colon cancers harbor different clinical and biologic characteristics. Right-sided colon cancers are more likely to have genome-wide hypermethylation via the CpG island methylator phenotype (CIMP), hypermutated state via microsatellite instability, and BRAF mutation. There are also differential exposures to potential carcinogenic toxins and microbiota in the right and left colon. Gene expression analyses further shed light on distinct biologic subtypes of colorectal cancers (CRCs), with 4 consensus molecular subtypes (CMSs) identified. Importantly, these subtypes are differentially distributed between right- and left-sided CRCs, with greater proportions of the "microsatellite unstable/immune" CMS1 and the "metabolic" CMS3 subtypes found in right-sided colon cancers. This review summarizes important biologic distinctions between right- and left-sided CRCs that likely impact prognosis and may predict for differential responses to biologic therapy. Given the inferior prognosis of stage III-IV right-sided CRCs and emerging data suggesting that anti-epidermal growth factor receptor antibody therapy is associated with worse survival in right-sided stage IV CRCs compared with left-sided cancers, these biologic differences between right- and left-sided CRCs provide critical context and may provide opportunities to personalize therapy. Copyright © 2017 by the National Comprehensive Cancer Network.

  11. Cardiac MRI in patients with complex CHD following primary or secondary implantation of MRI-conditional pacemaker system.

    Science.gov (United States)

    Al-Wakeel, Nadya; O h-Ici, Darach; Schmitt, Katharina R; Messroghli, Daniel R; Riesenkampff, Eugénie; Berger, Felix; Kuehne, Titus; Peters, Bjoern

    2016-02-01

    In patients with CHD, cardiac MRI is often indicated for functional and anatomical assessment. With the recent introduction of MRI-conditional pacemaker systems, cardiac MRI has become accessible for patients with pacemakers. The present clinical study aims to evaluate safety, susceptibility artefacts, and image reading of cardiac MRI in patients with CHD and MRI-conditional pacemaker systems. Material and methods CHD patients with MRI-conditional pacemaker systems and a clinical need for cardiac MRI were examined with a 1.5-T MRI system. Lead function was tested before and after MRI. Artefacts and image readings were evaluated using a four-point grading scale. A total of nine patients with CHD (mean age 34.0 years, range 19.5-53.6 years) received a total of 11 cardiac MRI examinations. Owing to clinical indications, seven patients had previously been converted from conventional to MRI-conditional pacemaker systems. All MRI examinations were completed without adverse effects. Device testing immediately after MRI and at follow-up showed no alteration of pacemaker device and lead function. Clinical questions could be addressed and answered in all patients. Cardiac MRI can be performed safely with high certainty of diagnosis in CHD patients with MRI-conditional pacemaker systems. In case of clinically indicated lead and box changing, CHD patients with non-MRI-conditional pacemaker systems should be considered for complete conversion to MRI-conditional systems.

  12. Electroconvulsive therapy in patients with cardiac pacemakers and implantable cardioverter defibrillators.

    Science.gov (United States)

    Dolenc, Tamara J; Barnes, Roxann D; Hayes, David L; Rasmussen, Keith G

    2004-09-01

    Electroconvulsive therapy (ECT) is used to treat major depressive illness, especially in elderly and medically frail patients. Not uncommonly, these patients have cardiac pacemakers or implantable cardioverter defibrillators (ICDs). Only a few case reports in the literature describe the use of ECT in such patients. Herein we review our ECT experience treating 26 pacemaker patients and 3 ICD patients. All patients obtained significant antidepressant benefits with ETC. Only one serious cardiac event occurred, a case of supraventricular tachycardia (SVT) requiring a stay on the cardiac intensive care unit. The SVT resolved and the patient went on to receive further uncomplicated ECT treatments. We conclude from this experience that with proper pre-ECT cardiac and pacemaker/defibrillator assessment, ECT can be safely and effectively administered to patients with an implanted cardiac device.

  13. Massive Left-sided Congestive Colitis Due to Idiopathic Inferior Mesenteric Arteriovenous Malformation

    Directory of Open Access Journals (Sweden)

    Laura Martí Gelonch

    2017-01-01

    Full Text Available Arteriovenous malformations (AVM of the inferior mesenteric artery are rare. They may be primary (congenital or idiopathic or secondary (acquired after trauma or of iatrogenic origin. Of the abdominal AVM, the inferior mesenteric trunk is the least commonly involved. Most reported cases are of iatrogenic origin, resulting from colon surgery. Only 17 cases have been described and published in the literature. The objective of this work is to make known a case treated in our center. We present a case of 73-year old male, who came to the emergency service with symptoms of abdominal distension, pain lasting 48 hours along with months of diarrhoea. CT scan and an abdominal CT angiography showed a massive left-sided congestive colitis due to idiopathic inferior mesenteric arteriovenous malformation. In our case, the decision was to carry out the treatment in two stages. Embolisation was performed in the first stage in order to decrease the blood flow and the risk of intraoperative bleeding, followed by resective surgery of the affected colon.

  14. Incidence of Bradycardia and Outcomes of Patients Who Underwent Orbital Atherectomy Without a Temporary Pacemaker.

    Science.gov (United States)

    Lee, Michael S; Nguyen, Heajung; Shlofmitz, Richard

    2017-02-01

    We analyzed the incidence of bradycardia and the safety of patients with severely calcified coronary lesions who underwent orbital atherectomy without the insertion of a temporary pacemaker. The presence of severely calcified coronary lesions can increase the complexity of percutaneous coronary intervention due to the difficulty in advancing and optimally expanding the stent. High-pressure inflations to predilate calcified lesions may cause angiographic complications like perforation and dissection. Suboptimal stent expansion is associated with stent thrombosis and restenosis. Orbital atherectomy safely and effectively modifies calcified plaque to facilitate optimal stent expansion. The incidence of bradycardia in orbital atherectomy is unknown. Fifty consecutive patients underwent orbital atherectomy from February 2014 to September 2016 at our institution, none of whom underwent insertion of a temporary pacemaker. The final analysis included 47 patients in this retrospective study as 3 patients were excluded because of permanent pacemaker implantation. The primary endpoint was significant bradycardia, defined as bradycardia requiring emergent pacemaker placement or a heart rate pacemaker appears to be safe.

  15. The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators

    International Nuclear Information System (INIS)

    Venselaar, J.L.M.

    1985-01-01

    Eight cardiac pacemakers were irradiated in a cobalt-60 beam. Two out of six demand-type pacemakers showed an alarming decrease in pulse repetition frequency when irradiated to dose levels that are used in radiotherapy. Two modern programmable pacemakers showed a failure at a dose of 97 and 147 Gy, respectively. The dose levels at which these failures occurred were low enough to recommend that cardiac pacemakers should always be kept outside the radiation beam. The signals induced by electromagnetic interference (EMI) from two linear accelerators were measured using a simulation model of a pacemaker. In the laboratory, 22 modern-type pacemakers were tested with these signals to determine the sensitivity for the electromagnetic fields in the treatment rooms. It was observed that an inhibition of one pacemaker pulse was to be expected on one of the two linear accelerators when switching the machine on and off. No permanent effects were found. These findings resulted in the recommendation in our department not to use this treatment machine for radiation therapy of pacemaker-bearing patients. (orig.)

  16. Application of RI power sources to cardiac pacemakers and aftercare in its implantation

    International Nuclear Information System (INIS)

    Hori, Motokazu

    1974-01-01

    RI power sources have long life when they are implanted into human bodies together with cardiac pacemakers, as compared with e.g. mercury batteries. Therefore, the frequency of their replacement can be by far less. However, there are the problems of radiation protection, high cost, availability, etc. The following matters are described: The cardiac pacemaker and its power supply, implantation into human body, problems with patients and conventional power sources; the current state of RI power sources for cardiac pacemakers, including plutonium-238 RTG and 147 Pm and 3 H batteries; and problems with the RI power sources. (Mori, K.)

  17. Ways and means to check on a patient wearing a cardiac pacemaker

    International Nuclear Information System (INIS)

    Marco, J.; Couderc, J.-J.; Salvador, M.; Dardenne, P.

    1975-01-01

    The implantation of present-day pacemakers with improved mercury batteries, undoubtedly increasing life expectancy and safety, or of new nuclear or lithium-battery pacemakers, will change the check-up frequency and delay prophylactic replacement operations. Similarly the developing possibilities of the telephone check-up, which offers a reliable long-distance analysis of the different parameters stated, will simplify the organization of these specialized consultations which are a heavy burden for cardiological Services fitting pacemakers and oblige patients to make special journeys, sometimes difficult and even impossible in certain cases. However a knowledge of these different factors by all practising doctors and cardiologists now seems indispensable for the safety of these increasingly numerous patients [fr

  18. Acute resection versus bridge to surgery with diverting colostomy for patients with acute malignant left sided colonic obstruction : Systematic review and meta-analysis

    NARCIS (Netherlands)

    Amelung, Femke J.; Mulder, C.L.J.; Verheijen, P.M.; Draaisma, W. A.; Siersema, P.D.; Consten, E. C J

    2015-01-01

    Background Currently, no consensus exists on the best treatment strategy for acute malignant left-sided colonic obstruction. This systematic review and meta-analysis aims to compare the outcomes following the two surgical treatment options; primary resection versus colostomy creation as bridge to

  19. Acute resection versus bridge to surgery with diverting colostomy for patients with acute malignant left sided colonic obstruction: Systematic review and meta-analysis

    NARCIS (Netherlands)

    Amelung, F.J.; Mulder, C.L.; Verheijen, P.M.; Draaisma, W.A.; Siersema, P.D.; Consten, E.C.

    2015-01-01

    BACKGROUND: Currently, no consensus exists on the best treatment strategy for acute malignant left-sided colonic obstruction. This systematic review and meta-analysis aims to compare the outcomes following the two surgical treatment options; primary resection versus colostomy creation as bridge to

  20. Inversion of the radionuclide regurgitant index in right-sided valvular regurgitation

    Energy Technology Data Exchange (ETDEWEB)

    Novack, H.; Machac, J.; Horowitz, S.F.

    1985-11-01

    Estimation of left-sided valvular insufficiency has been obtained using the ratio of left- to right-ventricular stroke counts, i.e., the regurgitant index. The present study was designed to evaluate the usefulness of the regurgitant index in identifying patients with isolated right-sided valvular insufficiency. We identified 12 patients with tricuspid or pulmonic regurgitation by at least two of the following criteria: (1) pulsatile liver, (2) positive Carvallo's sign, and (3) pulsatile jugular-venous distension. In 9 of the 12 patients, the right-sided insufficiency was confirmed by catheterization or contrast echocardiography and flow-directed pulsed-echo Doppler. The regurgitant index in patients with right-sided insufficiency was 0.59 +- 0.23. This was significantly different from patients with left-sided insufficiency (3.09 +- 0.8) and from control subjects (1.49 +- 0.32). In 11 of the 12 patients with right-sided regurgitant lesions, the regurgitant index was less than 1.0. The hepatic expansion fraction, a possible correlate of an expansile liver, has previously been found to be both sensitive and specific for the detection of patients with right-sided regurgitation. We calculated the hepatic expansion fraction in 6 patients with tricuspid regurgitation (including 3 with pulsatile livers) and 5 controls using the method of Handler et al.. In the present study, the hepatic expansion fraction in tricuspid-insufficiency patients was 4.3% as compared to 4.1% in normals (P=NS). In summary, this study suggests that the regurgitant index may be a sensitive tool for the diagnosis of right-sided regurgitant lesions, while the hepatic expansion fraction does not appear to be useful for identifying tricuspid insufficiency.

  1. Inversion of the radionuclide regurgitant index in right-sided valvular regurgitation

    International Nuclear Information System (INIS)

    Novack, H.; Machac, J.; Horowitz, S.F.; Mount Sinai Medical Center, New York

    1985-01-01

    Estimation of left-sided valvular insufficiency has been obtained using the ratio of left- to right-ventricular stroke counts, i.e., the regurgitant index. The present study was designed to evaluate the usefulness of the regurgitant index in identifying patients with isolated right-sided valvular insufficiency. We identified 12 patients with tricuspid or pulmonic regurgitation by at least two of the following criteria: (1) pulsatile liver, (2) positive Carvallo's sign, and (3) pulsatile jugular-venous distension. In 9 of the 12 patients, the right-sided insufficiency was confirmed by catheterization or contrast echocardiography and flow-directed pulsed-echo Doppler. The regurgitant index in patients with right-sided insufficiency was 0.59+-0.23. This was significantly different from patients with left-sided insufficiency (3.09+-0.8; P<0.001) and from control subjects (1.49+-0.32; P<0.001). In 11 of the 12 patients with right-sided regurgitant lesions, the regurgitant index was less than 1.0. The hepatic expansion fraction, a possible correlate of an expansile liver, has previously been found to be both sensitive and specific for the detection of patients with right-sided regurgitation. We calculated the hepatic expansion fraction in 6 patients with tricuspid regurgitation (including 3 with pulsatile livers) and 5 controls using the method of Handler et al.. In the present study, the hepatic expansion fraction in tricuspid-insufficiency patients was 4.3% as compared to 4.1% in normals (P=NS). In summary, this study suggests that the regurgitant index may be a sensitive tool for the diagnosis of right-sided regurgitant lesions, while the hepatic expansion fraction does not appear to be useful for identifying tricuspid insufficiency. (orig.)

  2. Evaluation of Safety and Efficacy of Qinming8631 DR Implantable Cardiac Pacemaker in Chinese Patients: A Prospective, Multicenter, Randomized Controlled Trial of the First Domestically Developed Pacemaker of China.

    Science.gov (United States)

    Xiang, Mei-Xiang; Wang, Dong-Qi; Xu, Jing; Zhang, Zheng; Hu, Jian-Xin; Wang, Dong-Mei; Gu, Xiang; Liu, He-Ping; Guo, Tao; Yang, Xiang-Jun; Ling, Feng; Lin, Jia-Feng; Cai, Shang-Lang; Zhu, Guo-Bin; Wang, Jian-An

    2016-11-20

    High cost of imported pacemakers is a main obstacle for Chinese patients suffering from bradyarrhythmia, and a domestically developed pacemaker will help lower the burden. This study aimed to evaluate the safety and efficacy of Qinming8631 DR (Qinming Medical, Baoji, China), the first domestically developed dual-chamber pacemaker of China, compared with a commercially available pacemaker Talos DR (Biotronik, Berlin, Germany) in Chinese patients. A prospective randomized trial was conducted at 14 centers in China. Participants were randomized into trial (Qinming8631 DR) and control (Talos DR) groups. Parameters of the pacing systems were collected immediately after device implantation and during follow-ups. The effective pacing rate at 6-month follow-up was recorded as the primary end point. Electrical properties, magnet response, single- and double-pole polarity conversion, rate response function, and adverse events of the pacing system were analyzed. The Cochran-Mantel-Haenszel Chi-square test, paired t-test, and Wilcoxon signed-rank test were used for measuring primary qualitative outcomes and comparing normally and abnormally distributed measurement data. A total of 225 patients with a diagnosis of bradyarrhythmia and eligible for this study were randomly enrolled into the trial (n = 113) and control (n = 112) groups. They underwent successful pacemaker implantation with acceptable postoperative pacing threshold and sensitivity. Effective pacing rates of trial and control groups were comparable both in the full analysis set and the per protocol set (81.4% vs. 79.5%, P = 0.712 and 95.4% vs. 89.5%, P = 0.143, respectively). In both data sets, noninferiority of the trial group was above the predefined noninferiority limit(-9.5%). This study established the noninferiority of Qinming8631 DR to Talos DR. The safety and efficacy of Qinming8631 DR pacemaker were comparable to those of Talos DR in treating patients with cardiac bradyarrhythmia.

  3. Handedness and dominant side of symptoms in Parkinson's disease.

    Science.gov (United States)

    Shi, Jie; Liu, Jie; Qu, Qiumin

    2014-02-20

    To investigate the association between handedness and the side of symptom dominance in Parkinson's disease (PD). One hundred and forty-six PD patients with symmetric symptoms (92 males and 54 females), aged 64.3 ± 9.1 years old, from a series of 247 PD patients were assessed for handedness and clinical features. The severity of PD was scored by unified Parkinson's disease rating scale (UPDRS) and Hoehn-Yahr staging on the "ON" state. Of 134 right-handed patients (91.8%), 83 (61.7%) had an initial onset on the right side (P=0.008), while of 12 left-handed patients (8.2%), 9 (75.0%) had an initial onset on the left side (P=0.013). Out of right-handed patients, 103 (76.9%) had the right-side dominance of PD symptoms (P<0.001). Among the left-handed subjects, 7 patients (58.3%) had left-sided and 5 patients (41.7%) had right-sided symptom dominance (P=0.564). In general, dominant side of symptoms was in accordance with handedness (P=0.008). In right-handed patients, rest tremor was the most common initial symptom (P<0.001), while rest tremor and rigidity-bradykinesia were initial symptoms in left-handed patients (P=0.366). PD symptoms emerge more often on the dominant hand-side, and the dominant side of symptoms is in accordance with handedness. Copyright © 2012 Elsevier España, S.L. All rights reserved.

  4. Left versus right deceased donor renal allograft outcome.

    LENUS (Irish Health Repository)

    Phelan, Paul J

    2009-12-01

    It has been suggested that the left kidney is easier to transplant than the right kidney because of the longer length of the left renal vein, facilitating the formation of the venous anastomosis. There are conflicting reports of differing renal allograft outcomes based on the side of donor kidney transplanted (left or right).We sought to determine the effect of side of donor kidney on early and late allograft outcome in our renal transplant population. We performed a retrospective analysis of transplanted left-right deceased donor kidney pairs in Ireland between January 1, 1998 and December 31, 2008. We used a time to death-censored graft failure approach for long-term allograft survival and also examined serum creatinine at different time points post-transplantation. All outcomes were included from day of transplant onwards. A total of 646 transplants were performed from 323 donors. The incidence of delayed graft function was 16.1% in both groups and there was no significant difference in acute rejection episodes or serum creatinine from 1 month to 8 years post-transplantation.There were 47 death-censored allograft failures in the left-sided group compared to 57 in the right-sided group (P = 0.24). These observations show no difference in renal transplant outcome between the recipients of left- and right-sided deceased donor kidneys.

  5. Analysis of a five year experience of permanent pacemaker implantation at a Nigerian Teaching Hospital: need for a national database

    Science.gov (United States)

    Falase, Bode; Sanusi, Michael; Johnson, Adeyemi; Akinrinlola, Fola; Ajayi, Reina; Oke, David

    2013-01-01

    Introduction Permanent pacemaker implantation is available in Nigeria. There is however no national registry or framework for pacemaker data collection. A pacemaker database has been developed in our institution and the results are analyzed in this study. Methods The study period was between January 2008 and December 2012. Patient data was extracted from a prospectively maintained database which was designed to include the fields of the European pacemaker patient identification code. Results Of the 51 pacemaker implants done, there were 29 males (56.9%) and 22 females (43.1%). Mean age was 68.2±12.7 years. Clinical indications were syncopal attacks in 25 patients (49%), dizzy spells in 15 patients (29.4%), bradycardia with no symptoms in 10 patients (17.7%) and dyspnoea in 2 patients (3.9%). The ECG diagnosis was complete heart block in 27 patients (53%), second degree heart block in 19 patients (37.2%) and sick sinus syndrome with bradycardia in 5 patients (9.8%). Pacemaker modes used were ventricular pacing in 29 patients (56.9%) and dual chamber pacing in 22 patients (43.1%). Files have been closed in 20 patients (39.2%) and 31 patients (60.8%) are still being followed up with median follow up of 26 months, median of 5 visits and 282 pacemaker checks done. Complications seen during follow up were 3 lead displacements (5.9%), 3 pacemaker infections (5.9%), 2 pacemaker pocket erosions (3.9%), and 1 pacemaker related death (2%). There were 5 non-pacemaker related deaths (9.8%). Conclusion Pacemaker data has been maintained for 5 years. We urge other implanting institutions in Nigeria to maintain similar databases and work towards establishment of a national pacemaker registry. PMID:24498465

  6. Short- and long-term need for permanent pacemaker after transcatheter implantation of the Edwards Sapien aortic valve prosthesis.

    Science.gov (United States)

    Moreno, Raúl; Calvo, Luis; Sánchez-Recalde, Angel; Galeote, Guillermo; Jiménez-Valero, Santiago; López, Teresa; Plaza, Ignacio; González-Davia, Rosa; Ramírez, Ulises; Mesa, Jose Maria; Moreno-Gomez, Isidro; López-Sendón, José-Luis

    2015-11-01

    A permanent pacemaker is frequently needed after transcatheter aortic valve implantation, but the available data are mainly on the CoreValve system. To evaluate the need for new permanent pacemaker after implantation of the Edwards Sapien device, as well as related factors. We included the first 100 patients treated with the Edwards Sapien device at our institution. Of these, 12 had a permanent pacemaker before the procedure, and thus our study population was the remaining 88 patients. A permanent pacemaker was indicated in eight patients (9.1%) during hospitalization or at 30 days. After discharge, another four patients needed a pacemaker (at 42 days and three, 18, and 30 months). Two variables were associated with the need for pacemaker during hospitalization: previous dialysis (13% vs. 1%, p=0.042) and complete right bundle branch block before the procedure (25% vs. 5%, p=0.032). More than one month after the procedure, the characteristics associated with the need for pacemaker were plasma creatinine level (2.5±1.7 vs. 1.3±0.6 mg/dl, p=0.001) and previous myocardial infarction (50% vs. 10%, p=0.013). The rate of pacemaker implantation with the Edwards Sapien device was 9.1%. Right bundle branch block and dialysis were associated with this complication.

  7. Intensity modulated radiotherapy with fixed collimator jaws for locoregional left-sided breast cancer irradiation.

    Science.gov (United States)

    Wang, Juanqi; Yang, Zhaozhi; Hu, Weigang; Chen, Zhi; Yu, Xiaoli; Guo, Xiaomao

    2017-05-16

    The purpose of this study is to evaluate the intensity modulated radiotherapy (IMRT) with the fixed collimator jaws technique (FJT) for the left breast and regional lymph node. The targeted breast tissue and the lymph nodes, and the normal tissues were contoured for 16 left-sided breast cancer patients previously treated with radiotherapy after lumpectomy. For each patient, treatment plans using different planning techniques, i.e., volumetric modulated arc therapy (VMAT), tangential IMRT (tangential-IMRT), and IMRT with FJT (FJT-IMRT) were developed for dosimetric comparisons. A dose of 50Gy was prescribed to the planning target volume. The dose-volume histograms were generated, and the paired t-test was used to analyze the dose differences. FJT-IMRT had similar mean heart volume receiving 30Gy (V30 Gy) with tangential-IMRT (1.5% and 1.6%, p = 0.41), but inferior to the VMAT (0.8%, p < 0.001). In the average heart mean dose comparison, FJT-IMRT had the lowest value, and it was 0.6Gy lower than that for the VMAT plans (p < 0.01). A significant dose increase in the contralateral breast and lung was observed in VMAT plans. Compared with tangential-IMRT and VMAT plans, FJT-IMRT reduced the mean dose of thyroid, humeral head and cervical esophageal by 47.6% (p < 0.01) and 45.7% (p < 0.01), 74.3% (p =< 0.01) and 73% (p =< 0.01), and 26.7% (p =< 0.01) and 29.2% (p =< 0.01). In conclusion, compared with tangential-IMRT and VMAT, FJT-IMRT plan has the lowest thyroid, humeral head and cervical esophageal mean dose and it can be a reasonable treatment option for a certain subgroup of patients, such as young left-breast cancer patients and/or patients with previous thyroid disease.

  8. Recovery of Ventriculo-Atrial Conduction after Adrenaline in Patients Implanted with Pacemakers.

    Science.gov (United States)

    Cismaru, Gabriel; Gusetu, Gabriel; Muresan, Lucian; Rosu, Radu; Andronache, Marius; Matuz, Roxana; Puiu, Mihai; Mester, Petru; Miclaus, Maria; Pop, Dana; Mircea, Petru Adrian; Zdrenghea, Dumitru

    2015-07-01

    Ventriculo-atrial (VA) conduction can have negative consequences for patients with implanted pacemakers and defibrillators. There is concern whether impaired VA conduction could recover during stressful situations. Although the influence of isoproterenol and atropine are well established, the effect of adrenaline has not been studied systematically. The objective of this study was to determine if adrenaline can facilitate recovery of VA conduction in patients implanted with pacemakers. A prospective study was conducted on 61 consecutive patients during a 4-month period (April-July 2014). The presence of VA conduction was assessed during the pacemaker implantation procedure. In case of an impaired VA conduction, adrenaline infusio was used as a stress surrogate to test conduction recovery. The indications for pacemaker implantation were: sinus node dysfunction in 18 patients, atrioventricular (AV) block in 40 patients, binodal dysfunction (sinus node+ AV node) in two patients and other (carotid sinus syndrome) in one patient. In the basal state, 15/61 (24.6%) presented spontaneous VA conduction and 46/61 (75.4%) had no VA conduction. After administration of adrenaline, there was VA conduction recovery in 5/46 (10.9%) patients. Adrenaline infusion produced recovery of VA conduction in 10.9% of patients with absent VA conduction in a basal state. Recovery of VA conduction during physiological or pathological stresses could be responsible for the pacemaker syndrome, PMT episodes, or certain implantable cardiac defibrillator detection issues. © 2015 Wiley Periodicals, Inc.

  9. Placement Of Cardiac PacemaKEr Trial (POCKET – rationale and design: a randomized controlled trial

    Directory of Open Access Journals (Sweden)

    Peter Magnusson

    2017-04-01

    Full Text Available BackgroundA pacemaker system consists of one or two leads connected to a device that is implanted into a pocket formed just below the collarbone. This pocket is typically subcutaneous, that is, located just above the pectoral fascia. Even though the size of pacemakers has decreased markedly, complications due to superficial implants do occur. An alternative technique would be intramuscular placement of the pacemaker device, but there are no randomized controlled trials (RCTs to support this approach, which is the rationale for the Placement Of Cardiac PacemaKEr Trial (POCKET. The aim is to study if intramuscular is superior to subcutaneous placement of a pacemaker pocket.MethodsIn October 2016, we started to enroll 200 consecutive patients with an indication for bradycardia pacemaker implantation. Patients are randomized to random block sizes, stratified by age group (cut-off: 65 years and sex, and then randomized to either subcutaneous or intramuscular implant. A concealed allocation procedure is employed, using sequentially numbered, sealed envelopes. Pocket site is blinded to the patient and in all subsequent care. The primary endpoint is patient overall satisfaction with the pocket location at 24 months as measured using a visual analog scale (VAS 0-10. Secondary endpoints are: complications, patient-reported satisfaction at 1, 12, and 24 months (overall satisfaction, pain, discomfort, degree of unsightly appearance, movement problems, and sleep problems due to device.ConclusionsPOCKET is a prospective interventional RCT designed to evaluate if intramuscular is superior to subcutaneous placement of a bradycardia pacemaker during a two-year follow-up.

  10. SU-E-T-585: Optically-Stimulated Luminescent Dosimeters for Monitoring Pacemaker Dose in Radiation Therapy

    International Nuclear Information System (INIS)

    Apicello, L; Riegel, A; Jamshidi, A

    2015-01-01

    Purpose: A sufficient amount of ionizing radiation can cause failure to components of pacemakers. Studies have shown that permanent damage can occur after a dose of 10 Gy and minor damage to functionality occurs at doses as low as 2 Gy. Optically stimulated thermoluminescent dosimeters (OSLDs) can be used as in vivo dosimeters to predict dose to be deposited throughout the treatment. The purpose of this work is to determine the effectiveness of using OSLDs for in vivo dosimetry of pacemaker dose. Methods: As part of a clinical in vivo dosimetry experience, OSLDs were placed at the site of the pacemaker by the therapist for one fraction of the radiation treatment. OSLD measurements were extrapolated to the total dose to be received by the pacemaker during treatment. A total of 79 measurements were collected from November 2011 to December 2013 on six linacs. Sixty-six (66) patients treated in various anatomical sites had the dose of their pacemakers monitored. Results: Of the 79 measurements recorded, 76 measurements (96 %) were below 2 Gy. The mean and standard deviation were 50.12 ± 76.41 cGy. Of the 3 measurements that exceeded 2 Gy, 2 measurements matched the dose predicted in the treatment plan and 1 was repeated after an unexpectedly high Result. The repeated measurement yielded a total dose less than 2 Gy. Conclusion: This analysis suggests OSLDs may be used for in vivo monitoring of pacemaker dose. Further research should be performed to assess the effect of increased backscatter from the pacemaker device

  11. Placement Of Cardiac PacemaKEr Trial (POCKET – rationale and design: a randomized controlled trial

    Directory of Open Access Journals (Sweden)

    Peter Magnusson

    2017-04-01

    Full Text Available Background: A pacemaker system consists of one or two leads connected to a device that is implanted into a pocket formed just below the collarbone. This pocket is typically subcutaneous, that is, located just above the pectoral fascia. Even though the size of pacemakers has decreased markedly, complications due to superficial implants do occur. An alternative technique would be intramuscular placement of the pacemaker device, but there are no randomized controlled trials (RCTs to support this approach, which is the rationale for the Placement Of Cardiac PacemaKEr Trial (POCKET. The aim is to study if intramuscular is superior to subcutaneous placement of a pacemaker pocket. Methods: In October 2016, we started to enroll 200 consecutive patients with an indication for bradycardia pacemaker implantation. Patients are randomized to random block sizes, stratified by age group (cut-off: 65 years and sex, and then randomized to either subcutaneous or intramuscular implant. A concealed allocation procedure is employed, using sequentially numbered, sealed envelopes. Pocket site is blinded to the patient and in all subsequent care. The primary endpoint is patient overall satisfaction with the pocket location at 24 months as measured using a visual analog scale (VAS 0-10. Secondary endpoints are: complications, patient-reported satisfaction at 1, 12, and 24 months (overall satisfaction, pain, discomfort, degree of unsightly appearance, movement problems, and sleep problems due to device. Conclusions: POCKET is a prospective interventional RCT designed to evaluate if intramuscular is superior to subcutaneous placement of a bradycardia pacemaker during a two-year follow-up.

  12. SU-E-T-585: Optically-Stimulated Luminescent Dosimeters for Monitoring Pacemaker Dose in Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Apicello, L [Hofstra University, Hempstead, NY (United States); Riegel, A; Jamshidi, A [North Shore LIJ Health System, Lake Success, NY (United States)

    2015-06-15

    Purpose: A sufficient amount of ionizing radiation can cause failure to components of pacemakers. Studies have shown that permanent damage can occur after a dose of 10 Gy and minor damage to functionality occurs at doses as low as 2 Gy. Optically stimulated thermoluminescent dosimeters (OSLDs) can be used as in vivo dosimeters to predict dose to be deposited throughout the treatment. The purpose of this work is to determine the effectiveness of using OSLDs for in vivo dosimetry of pacemaker dose. Methods: As part of a clinical in vivo dosimetry experience, OSLDs were placed at the site of the pacemaker by the therapist for one fraction of the radiation treatment. OSLD measurements were extrapolated to the total dose to be received by the pacemaker during treatment. A total of 79 measurements were collected from November 2011 to December 2013 on six linacs. Sixty-six (66) patients treated in various anatomical sites had the dose of their pacemakers monitored. Results: Of the 79 measurements recorded, 76 measurements (96 %) were below 2 Gy. The mean and standard deviation were 50.12 ± 76.41 cGy. Of the 3 measurements that exceeded 2 Gy, 2 measurements matched the dose predicted in the treatment plan and 1 was repeated after an unexpectedly high Result. The repeated measurement yielded a total dose less than 2 Gy. Conclusion: This analysis suggests OSLDs may be used for in vivo monitoring of pacemaker dose. Further research should be performed to assess the effect of increased backscatter from the pacemaker device.

  13. Hypoplastic left heart syndrome

    Directory of Open Access Journals (Sweden)

    Thiagarajan Ravi

    2007-05-01

    Full Text Available Abstract Hypoplastic left heart syndrome(HLHS refers to the abnormal development of the left-sided cardiac structures, resulting in obstruction to blood flow from the left ventricular outflow tract. In addition, the syndrome includes underdevelopment of the left ventricle, aorta, and aortic arch, as well as mitral atresia or stenosis. HLHS has been reported to occur in approximately 0.016 to 0.036% of all live births. Newborn infants with the condition generally are born at full term and initially appear healthy. As the arterial duct closes, the systemic perfusion becomes decreased, resulting in hypoxemia, acidosis, and shock. Usually, no heart murmur, or a non-specific heart murmur, may be detected. The second heart sound is loud and single because of aortic atresia. Often the liver is enlarged secondary to congestive heart failure. The embryologic cause of the disease, as in the case of most congenital cardiac defects, is not fully known. The most useful diagnostic modality is the echocardiogram. The syndrome can be diagnosed by fetal echocardiography between 18 and 22 weeks of gestation. Differential diagnosis includes other left-sided obstructive lesions where the systemic circulation is dependent on ductal flow (critical aortic stenosis, coarctation of the aorta, interrupted aortic arch. Children with the syndrome require surgery as neonates, as they have duct-dependent systemic circulation. Currently, there are two major modalities, primary cardiac transplantation or a series of staged functionally univentricular palliations. The treatment chosen is dependent on the preference of the institution, its experience, and also preference. Although survival following initial surgical intervention has improved significantly over the last 20 years, significant mortality and morbidity are present for both surgical strategies. As a result pediatric cardiologists continue to be challenged by discussions with families regarding initial decision

  14. Safety of holmium laser prostatectomy in patients with cardiac pacemaker implant

    Directory of Open Access Journals (Sweden)

    Narmada P Gupta

    2006-01-01

    Full Text Available Objectives: The use of the standard monopolar electrocautery is associated with significant risks of implant malfunction in patients on a cardiac pacemaker. It is also associated with a risk of adverse cardiac events due to blood loss and fluid absorption. The properties of the holmium laser prevent the occurrence of these adverse events. We report the successful use of this technology in resecting the gland in patients on a permanent cardiac pacemaker implant. MATERIALS AND Methods: Six patients with permanent cardiac pacemaker implant were treated with holmium laser resection of prostate over a period of two years. Treated patients had bothersome prostatic symptoms and failed to respond to medical therapy. All patients were operated under spinal anesthesia using a high power VersaPulse ® PowerSuiteTM Holmium laser source. Normal saline was used as irrigant. Intravesical tissue morcellator was also used to remove the larger fragments in two of the patients. Results : Median patient age was 60 years (range 56-73 and median prostate volume was 40cc (range 20-48cc. None of the patient required blood transfusion or had significant hyponatremia or Transurethral resection syndrome. No patients had any pacemaker malfunction or hemodynamic instability during the procedure or in immediate postoperative period. Improvement in maximum urine flow rate was observed from an average of 7 ml/sec in preoperative period to 22 ml/sec postoperatively at 3 month followup. Conclusions: Holmium laser prostatectomy offers the ideal modality of surgery in patients on a cardiac pacemaker. It helps to avoid additional preparation and minimizes the risk of device malfunction and adverse post operative events.

  15. Left-sided breast cancer irradiation using rotational and fixed-field radiotherapy

    International Nuclear Information System (INIS)

    Qi, X. Sharon; Liu, Tian X.; Liu, Arthur K.; Newman, Francis; Rabinovitch, Rachel; Kavanagh, Brian; Hu, Y. Angie

    2014-01-01

    The 3-dimensional conformal radiotherapy (3DCRT) technique is the standard for breast cancer radiotherapy. During treatment planning, not only the coverage of the planning target volume (PTV) but also the minimization of the dose to critical structures, such as the lung, heart, and contralateral breast tissue, need to be considered. Because of the complexity and variations of patient anatomy, more advanced radiotherapy techniques are sometimes desired to better meet the planning goals. In this study, we evaluated external-beam radiation treatment techniques for left breast cancer using various delivery platforms: fixed-field including TomoDirect (TD), static intensity-modulated radiotherapy (sIMRT), and rotational radiotherapy including Elekta volumetric-modulated arc therapy (VMAT) and tomotherapy helical (TH). A total of 10 patients with left-sided breast cancer who did or did not have positive lymph nodes and were previously treated with 3DCRT/sIMRT to the entire breast were selected, their treatment was planned with Monaco VMAT, TD, and TH. Dosimetric parameters including PTV coverage, organ-at-risk (OAR) sparing, dose-volume histograms, and target minimum/maximum/mean doses were evaluated. It is found that for plans providing comparable PTV coverage, the Elekta VMAT plans were generally more inhomogeneous than the TH and TD plans. For the cases with regional node involvement, the average mean doses administered to the heart were 9.2 (± 5.2) and 8.8 (± 3.0) Gy in the VMAT and TH plans compared with 11.9 (± 6.4) and 11.8 (± 9.2) Gy for the 3DCRT and TD plans, respectively, with slightly higher doses given to the contralateral lung or breast or both. On average, the total monitor units for VMAT plans are 11.6% of those TH plans. Our studies have shown that VMAT and TH plans offer certain dosimetric advantages over fixed-field IMRT plans for advanced breast cancer requiring regional nodal treatment. However, for early-stage breast cancer fixed

  16. Sleeping and resting respiratory rates in dogs and cats with medically-controlled left-sided congestive heart failure.

    Science.gov (United States)

    Porciello, F; Rishniw, M; Ljungvall, I; Ferasin, L; Haggstrom, J; Ohad, D G

    2016-01-01

    Sleeping and resting respiratory rates (SRR and RRR, respectively) are commonly used to monitor dogs and cats with left-sided cardiac disease and to identify animals with left-sided congestive heart failure (L-CHF). Dogs and cats with subclinical heart disease have SRRmean values dogs and cats with CHF that is well controlled with medical therapy. In this study, SRR and RRR were measured by the owners of 51 dogs and 22 cats with stable, well-controlled CHF. Median canine SRRmean was 20 breaths/min (7-39 breaths/min); eight dogs were ≥25 breaths/min and one dog only was ≥30 breaths/min. Canine SRRmean was unrelated to pulmonary hypertension or diuretic dose. Median feline SRRmean was 20 breaths/min (13-31 breaths/min); four cats were ≥25 breaths/min and only one cat was ≥30 breaths/min. Feline SRRmean was unrelated to diuretic dose. SRR remained stable during collection in both species with little day-to-day variability. The median canine RRRmean was 24 breaths/min (12-44 breaths/min), 17 were ≥25 breaths/min, seven were ≥30 breaths/min, two were >40 breaths/min. Median feline RRRmean was 24 breaths/min (15-45 breaths/min); five cats had RRRmean ≥25 breaths/min; one had ≥30 breaths/min, and two had ≥40 breaths/min. These data suggest that most dogs and cats with CHF that is medically well-controlled and stable have SRRmean and RRRmean dogs and cats. Copyright © 2015 Elsevier Ltd. All rights reserved.

  17. Incidence and predictors of asymptomatic atrial fibrillation in patients older than 70 years with complete atrioventricular block and dual chamber pacemaker implantation.

    LENUS (Irish Health Repository)

    Radeljic, Vjekoslav

    2012-01-31

    AIM: To evaluate predictors of asymptomatic atrial fibrillation in patients older than 70 years with complete atrioventricular (AV) block, normal left ventricular systolic function, and implanted dual chamber (DDD) pacemaker. METHODS: Hundred and eighty six patients with complete AV block were admitted over one year to the Sisters of Mercy University Hospital. The study recruited patients older than 70 years, with no history of atrial fibrillation, heart failure, or reduced left ventricular systolic function. All the patients were implanted with the same pacemaker. Out of 103 patients who were eligible for the study, 81 (78%) were evaluated. Among those 81 (78%) were evaluated. Eighty one (78%) patients were evaluated. Follow-up time ranged from 12 to 33 months (average +\\/-standard deviation 23 +\\/- 5 months). Primary end-point was asymptomatic atrial fibrillation occurrence recorded by the pacemaker. Atrial fibrillation occurrence was defined as atrial high rate episodes (AHRE) lasting >5 minutes. Binary logistic regression was used to identify the predictors of development of asymptomatic atrial fibrillation. Results. The 81 patients were stratified into two groups depending on the presence of AHRE lasting >5 minutes (group 1 had AHRE>5 minutes and group 2 AHRE<5 minutes). AHRE lasting >5 minutes were detected in 49 (60%) patients after 3 months and in 53 (65%) patients after 18 moths. After 3 months, only hypertension (odds ratio [OR], 17.63; P = 0.020) was identified as a predictor of asymptomatic atrial fibrillation. After 18 months, hypertension (OR, 14.0; P = 0.036), P wave duration >100 ms in 12 lead ECG (OR, 16.5; P = 0.001), and intracardial atrial electrogram signal amplitude >4 mV (OR, 4.27; P = 0.045) were identified as predictors of atrial fibrillation. CONCLUSION: In our study population, hypertension was the most robust and constant predictor of asymptomatic atrial fibrillation after 3 months, while P wave duration >100 ms in 12-lead ECG and

  18. Battery Malfunction of a Leadless Cardiac Pacemaker - A Worrisome Single-Center Experience.

    Science.gov (United States)

    Richter, Sergio; Döring, Michael; Ebert, Micaela; Bode, Kerstin; Müssigbrodt, Andreas; Sommer, Philipp; Husser, Daniela; Hindricks, Gerhard

    2018-03-14

    Leadless cardiac pacemaker (LCP) therapy has been established clinically as a feasible and safe alternative to conventional transvenous pacemaker therapy for patients with an indication for single-chamber right-ventricular pacing. 1-3 However, reports on loss of telemetry and pacing output due to abrupt battery failure called the safety of one of the two commercially available systems seriously into question. The initial battery advisory with the Nanostim™ LCP was issued by the manufacturer in October 2016, who instantly called a global stop to Nanostim™ implants. To this day, similar battery issues have not been described for the Micra™ transcatheter pacing system. Therefore, we thought to analyze the long-term pacemaker performance and rate of battery malfunction of the Nanostim™ LCP in our patient population.

  19. A rare localization in right-sided endocarditis diagnosed by echocardiography: A case report

    Directory of Open Access Journals (Sweden)

    Beaufils Philippe

    2003-08-01

    Full Text Available Abstract Background Right-sided endocarditis occurs predominantly in intravenous drug users, patients with pacemakers or central venous lines and with congenital heart diseases. The vast majority of cases involve the tricuspid valve. Case presentation A case of a 31-year-old woman with intravenous drug abuse who had a right-sided vegetation attached to the muscular bundle of the right ventricle is presented. Transthoracic echocardiography revealed a vegetation in the right ventricular outflow tract. Transesophageal echocardiography clearly showed that the 1.8 cm vegetation was not adherent to the pulmonary valve but attached to a muscular bundle. Conclusions Our case points to an unusual location of right-sided endocarditis in intravenous drug users. It confirms that TTE remains an easy and highly sensitive first-line examination for the diagnosis of right-sided endocarditis.

  20. Isolated persistent left superior vena cava: A case report and its clinical implications

    Directory of Open Access Journals (Sweden)

    Samarjit Bisoyi

    2017-01-01

    Full Text Available The venous anomaly of a persistent left superior vena cava (PLSVC affects 0.3%-0.5% of the general population. PLSVC with absent right superior vena cava, also termed as "isolated PLSVC," is an extremely rare venous anomaly. Almost half of the patients with isolated PLSVC have cardiac anomalies in the form of atrial septal defect, endocardial cushion defects, or tetralogy of Fallot. Isolated PLSVC is usually innocuous. Its discovery, however, has important clinical implications. It can pose clinical difficulties with central venous access, cardiothoracic surgeries, and pacemaker implantation. When it drains to the left atrium, it may create a right to left shunt. In this case report, we present the incidental finding of isolated PLSVC in a patient who underwent aortic valve replacement. Awareness about this condition and its variations is important to avoid complications.

  1. Pacemaker Use in New Zealand - Data From the New Zealand Implanted Cardiac Device Registry (ANZACS-QI 15).

    Science.gov (United States)

    Larsen, P D; Kerr, A J; Hood, M; Harding, S A; Hooks, D; Heaven, D; Lever, N A; Sinclair, S; Boddington, D; Tang, E W; Swampillai, J; Stiles, M K

    2017-03-01

    The New Zealand Cardiac Implanted Device Registry (Device) has recently been developed under the auspices of the New Zealand Branch of the Cardiac Society of Australia and New Zealand. This study describes the initial Device registry cohort of patients receiving a new pacemaker, their indications for pacing and their perioperative complications. The Device Registry was used to audit patients receiving a first pacemaker between 1 st January 2014 and 1 st June 2015. We examined 1611 patients undergoing first pacemaker implantation. Patients were predominantly male (59%), and had a median age of 70 years. The most common symptom for pacemaker implantation was syncope (39%), followed by dizziness (30%) and dyspnoea (12%). The most common aetiology for a pacemaker was a conduction tissue disorder (35%), followed by sinus node dysfunction (22%). Atrioventricular (AV) block was the most common ECG abnormality, present in 44%. Dual chamber pacemakers were most common (62%), followed by single chamber ventricular pacemakers (34%), and cardiac resynchronisation therapy - pacemakers (CRT-P) (2%). Complications within 24hours of the implant procedure were reported in 64 patients (3.9%), none of which were fatal. The most common complication was the need for reoperation to manipulate a lead, occurring in 23 patients (1.4%). This is the first description of data entered into the Device registry. Patients receiving a pacemaker were younger than in European registries, and there was a low use of CRT-P devices compared to international rates. Complications rates were low and compare favourably to available international data. Copyright © 2016 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

  2. Left preference for sport tasks does not necessarily indicate left-handedness: sport-specific lateral preferences, relationship with handedness and implications for laterality research in behavioural sciences.

    Directory of Open Access Journals (Sweden)

    Florian Loffing

    Full Text Available In the elite domain of interactive sports, athletes who demonstrate a left preference (e.g., holding a weapon with the left hand in fencing or boxing in a 'southpaw' stance seem overrepresented. Such excess indicates a performance advantage and was also interpreted as evidence in favour of frequency-dependent selection mechanisms to explain the maintenance of left-handedness in humans. To test for an overrepresentation, the incidence of athletes' lateral preferences is typically compared with an expected ratio of left- to right-handedness in the normal population. However, the normal population reference values did not always relate to the sport-specific tasks of interest, which may limit the validity of reports of an excess of 'left-oriented' athletes. Here we sought to determine lateral preferences for various sport-specific tasks (e.g., baseball batting, boxing in the normal population and to examine the relationship between these preferences and handedness. To this end, we asked 903 participants to indicate their lateral preferences for sport-specific and common tasks using a paper-based questionnaire. Lateral preferences varied considerably across the different sport tasks and we found high variation in the relationship between those preferences and handedness. In contrast to unimanual tasks (e.g., fencing or throwing, for bimanually controlled actions such as baseball batting, shooting in ice hockey or boxing the incidence of left preferences was considerably higher than expected from the proportion of left-handedness in the normal population and the relationship with handedness was relatively low. We conclude that (i task-specific reference values are mandatory for reliably testing for an excess of athletes with a left preference, (ii the term 'handedness' should be more cautiously used within the context of sport-related laterality research and (iii observation of lateral preferences in sports may be of limited suitability for the

  3. Left preference for sport tasks does not necessarily indicate left-handedness: sport-specific lateral preferences, relationship with handedness and implications for laterality research in behavioural sciences.

    Science.gov (United States)

    Loffing, Florian; Sölter, Florian; Hagemann, Norbert

    2014-01-01

    In the elite domain of interactive sports, athletes who demonstrate a left preference (e.g., holding a weapon with the left hand in fencing or boxing in a 'southpaw' stance) seem overrepresented. Such excess indicates a performance advantage and was also interpreted as evidence in favour of frequency-dependent selection mechanisms to explain the maintenance of left-handedness in humans. To test for an overrepresentation, the incidence of athletes' lateral preferences is typically compared with an expected ratio of left- to right-handedness in the normal population. However, the normal population reference values did not always relate to the sport-specific tasks of interest, which may limit the validity of reports of an excess of 'left-oriented' athletes. Here we sought to determine lateral preferences for various sport-specific tasks (e.g., baseball batting, boxing) in the normal population and to examine the relationship between these preferences and handedness. To this end, we asked 903 participants to indicate their lateral preferences for sport-specific and common tasks using a paper-based questionnaire. Lateral preferences varied considerably across the different sport tasks and we found high variation in the relationship between those preferences and handedness. In contrast to unimanual tasks (e.g., fencing or throwing), for bimanually controlled actions such as baseball batting, shooting in ice hockey or boxing the incidence of left preferences was considerably higher than expected from the proportion of left-handedness in the normal population and the relationship with handedness was relatively low. We conclude that (i) task-specific reference values are mandatory for reliably testing for an excess of athletes with a left preference, (ii) the term 'handedness' should be more cautiously used within the context of sport-related laterality research and (iii) observation of lateral preferences in sports may be of limited suitability for the verification of

  4. Left Preference for Sport Tasks Does Not Necessarily Indicate Left-Handedness: Sport-Specific Lateral Preferences, Relationship with Handedness and Implications for Laterality Research in Behavioural Sciences

    Science.gov (United States)

    Loffing, Florian; Sölter, Florian; Hagemann, Norbert

    2014-01-01

    In the elite domain of interactive sports, athletes who demonstrate a left preference (e.g., holding a weapon with the left hand in fencing or boxing in a ‘southpaw’ stance) seem overrepresented. Such excess indicates a performance advantage and was also interpreted as evidence in favour of frequency-dependent selection mechanisms to explain the maintenance of left-handedness in humans. To test for an overrepresentation, the incidence of athletes' lateral preferences is typically compared with an expected ratio of left- to right-handedness in the normal population. However, the normal population reference values did not always relate to the sport-specific tasks of interest, which may limit the validity of reports of an excess of ‘left-oriented’ athletes. Here we sought to determine lateral preferences for various sport-specific tasks (e.g., baseball batting, boxing) in the normal population and to examine the relationship between these preferences and handedness. To this end, we asked 903 participants to indicate their lateral preferences for sport-specific and common tasks using a paper-based questionnaire. Lateral preferences varied considerably across the different sport tasks and we found high variation in the relationship between those preferences and handedness. In contrast to unimanual tasks (e.g., fencing or throwing), for bimanually controlled actions such as baseball batting, shooting in ice hockey or boxing the incidence of left preferences was considerably higher than expected from the proportion of left-handedness in the normal population and the relationship with handedness was relatively low. We conclude that (i) task-specific reference values are mandatory for reliably testing for an excess of athletes with a left preference, (ii) the term ‘handedness’ should be more cautiously used within the context of sport-related laterality research and (iii) observation of lateral preferences in sports may be of limited suitability for the

  5. Pacemaker rate and depolarization block in nigral dopamine neurons: a somatic sodium channel balancing act

    Science.gov (United States)

    Tucker, Kristal R.; Huertas, Marco A.; Horn, John P.; Canavier, Carmen C.; Levitan, Edwin S.

    2012-01-01

    Midbrain dopamine (DA) neurons are slow intrinsic pacemakers that undergo depolarization (DP) block upon moderate stimulation. Understanding DP block is important because it has been correlated with the clinical efficacy of chronic antipsychotic drug treatment. Here we describe how voltage-gated sodium (NaV) channels regulate DP block and pacemaker activity in DA neurons of the substantia nigra using rat brain slices. The distribution, density and gating of NaV currents were manipulated by blocking native channels with tetrodotoxin and by creating virtual channels and anti-channels with dynamic clamp. Although action potentials initiate in the axon initial segment (AIS) and NaV channels are distributed in multiple dendrites, selective reduction of NaV channel activity in the soma was sufficient to decrease pacemaker frequency and increase susceptibility to DP block. Conversely, increasing somatic NaV current density raised pacemaker frequency and lowered susceptibility to DP block. Finally, when NaV currents were restricted to the soma, pacemaker activity occurred at abnormally high rates due to excessive local subthreshold NaV current. Together with computational simulations, these data show that both the slow pacemaker rate and the sensitivity to DP block that characterizes DA neurons result from the low density of somatic NaV channels. More generally, we conclude that the somatodendritic distribution of NaV channels is a major determinant of repetitive spiking frequency. PMID:23077037

  6. The influence of ionizing radiation on the probability of failure of long-term pacemakers

    International Nuclear Information System (INIS)

    Eger, H.; Hesse, W.; Otte, K.B.; Rauh, G.; Salewski, D.; Stopp, G.

    1982-01-01

    The functional disturbances observed during the exposure of cardiac pacemakers with a CMOS switching circle U 115 are described qualitatively for different types of radiation. Based on the maximum sensitivity of the U 115 within the X-ray range, the correlation for this range between dose of exposure and parameter changes is established. As is demonstrated such changes is established. As is demonstrated such changes are not clinically relevant as long as the dose of 5,16 x 10 -2 C/kg is not exceeded. Proceeding from the accumulated dose to which the U 115 is exposed during various diagnostic radiographic procedures it is shown that such procedures do not result in functional disturbances of the pacemaker. Recommendations are given for radiotherapeutic measures as to pacemaker patients, which should be realized in close co-operation between the pacemaker centre concerned and an authorized hospital. (author)

  7. Right Ear Advantage of Speech Audiometry in Single-sided Deafness.

    Science.gov (United States)

    Wettstein, Vincent G; Probst, Rudolf

    2018-04-01

    Postlingual single-sided deafness (SSD) is defined as normal hearing in one ear and severely impaired hearing in the other ear. A right ear advantage and dominance of the left hemisphere are well established findings in individuals with normal hearing and speech processing. Therefore, it seems plausible that a right ear advantage would exist in patients with SSD. The audiometric database was searched to identify patients with SSD. Results from the German monosyllabic Freiburg word test and four-syllabic number test in quiet were evaluated. Results of right-sided SSD were compared with left-sided SSD. Statistical calculations were done with the Mann-Whitney U test. Four hundred and six patients with SSD were identified, 182 with right-sided and 224 with left-sided SSD. The two groups had similar pure-tone thresholds without significant differences. All test parameters of speech audiometry had better values for right ears (SSD left) when compared with left ears (SSD right). Statistically significant results (p right and 97.5 ± 4.7% left, p right and 93.9 ± 9.1% left, p right and 63.8 ± 11.1 dB SPL left, p right ear advantage of speech audiometry was found in patients with SSD in this retrospective study of audiometric test results.

  8. Permanent pacemaker implantation in octogenarians with unexplained syncope and positive electrophysiologic testing.

    Science.gov (United States)

    Giannopoulos, Georgios; Kossyvakis, Charalampos; Panagopoulou, Vasiliki; Tsiachris, Dimitrios; Doudoumis, Konstantinos; Mavri, Maria; Vrachatis, Dimitrios; Letsas, Konstantinos; Efremidis, Michael; Katsivas, Apostolos; Lekakis, John; Deftereos, Spyridon

    2017-05-01

    Syncope is a common problem in the elderly, and a permanent pacemaker is a therapeutic option when a bradycardic etiology is revealed. However, the benefit of pacing when no association of symptoms to bradycardia has been shown is not clear, especially in the elderly. The aim of this study was to evaluate the effect of pacing on syncope-free mortality in patients aged 80 years or older with unexplained syncope and "positive" invasive electrophysiologic testing (EPT). This was an observational study. A positive EPT for the purposes of this study was defined by at least 1 of the following: a corrected sinus node recovery time of >525 ms, a basic HV interval of >55 ms, detection of infra-Hisian block, or appearance of second-degree atrioventricular block on atrial decremental pacing at a paced cycle length of >400 ms. Among the 2435 screened patients, 228 eligible patients were identified, 145 of whom were implanted with a pacemaker. Kaplan-Meier analysis determined that time to event (syncope or death) was 50.1 months (95% confidence interval 45.4-54.8 months) with a pacemaker vs 37.8 months (95% confidence interval 31.3-44.4 months) without a pacemaker (log-rank test, P = .001). The 4-year time-dependent estimate of the rate of syncope was 12% vs 44% (P pacemaker implantation was independently associated with longer syncope-free survival. Significant differences were also shown in the individual components of the primary outcome measure (syncope and death from any cause). Copyright © 2017 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

  9. A Parametric Computational Model of the Action Potential of Pacemaker Cells.

    Science.gov (United States)

    Ai, Weiwei; Patel, Nitish D; Roop, Partha S; Malik, Avinash; Andalam, Sidharta; Yip, Eugene; Allen, Nathan; Trew, Mark L

    2018-01-01

    A flexible, efficient, and verifiable pacemaker cell model is essential to the design of real-time virtual hearts that can be used for closed-loop validation of cardiac devices. A new parametric model of pacemaker action potential is developed to address this need. The action potential phases are modeled using hybrid automaton with one piecewise-linear continuous variable. The model can capture rate-dependent dynamics, such as action potential duration restitution, conduction velocity restitution, and overdrive suppression by incorporating nonlinear update functions. Simulated dynamics of the model compared well with previous models and clinical data. The results show that the parametric model can reproduce the electrophysiological dynamics of a variety of pacemaker cells, such as sinoatrial node, atrioventricular node, and the His-Purkinje system, under varying cardiac conditions. This is an important contribution toward closed-loop validation of cardiac devices using real-time heart models.

  10. Biventricular pacemaker implantation with the BV Pulsera

    Energy Technology Data Exchange (ETDEWEB)

    Wright, J.D.W.; Todd, D.; Waktare, J.E.P.; Hughes, S.; Abell, C. [The Cardiothoracic Centre, Liverpool (United Kingdom)

    2005-07-01

    The BV Pulsera mobile X-ray system is used for biventricular pacemaker implantation in a surgical environment. It provides high-quality images of the moving structures, including the fine coronary leads, and can be used for long fluoroscopy runs without overheating. (orig.)

  11. Biventricular pacemaker implantation with the BV Pulsera

    International Nuclear Information System (INIS)

    Wright, J.D.W.; Todd, D.; Waktare, J.E.P.; Hughes, S.; Abell, C.

    2005-01-01

    The BV Pulsera mobile X-ray system is used for biventricular pacemaker implantation in a surgical environment. It provides high-quality images of the moving structures, including the fine coronary leads, and can be used for long fluoroscopy runs without overheating. (orig.)

  12. Intra-cardiac pacemaker infection: Surgical management and outcome

    Directory of Open Access Journals (Sweden)

    Sameh Elameen

    2016-05-01

    Conclusion: Explantation of the complete pacemaker system has proved a reliable method to eradicate infection. Complications were rare, except in patients who present lately in a critically ill condition and septic shock.

  13. Colonic stenting as bridge to surgery versus emergency surgery for management of acute left-sided malignant colonic obstruction: a multicenter randomized trial (Stent-in 2 study

    Directory of Open Access Journals (Sweden)

    Scholten Pieter

    2007-07-01

    Full Text Available Abstract Background Acute left-sided colonic obstruction is most often caused by malignancy and the surgical treatment is associated with a high mortality and morbidity rate. Moreover, these operated patients end up with a temporary or permanent stoma. Initial insertion of an enteral stent to decompress the obstructed colon, allowing for surgery to be performed electively, is gaining popularity. In uncontrolled studies stent placement before elective surgery has been suggested to decrease mortality, morbidity and number of colostomies. However stent perforation can lead to peritoneal tumor spill, changing a potentially curable disease in an incurable one. Therefore it is of paramount importance to compare the outcomes of colonic stenting followed by elective surgery with emergency surgery for the management of acute left-sided malignant colonic obstruction in a randomized multicenter fashion. Methods/design Patients with acute left-sided malignant colonic obstruction eligible for this study will be randomized to either emergency surgery (current standard treatment or colonic stenting as bridge to elective surgery. Outcome measurements are effectiveness and costs of both strategies. Effectiveness will be evaluated in terms of quality of life, morbidity and mortality. Quality of life will be measured with standardized questionnaires (EORTC QLQ-C30, EORTC QLQ-CR38, EQ-5D and EQ-VAS. Morbidity is defined as every event leading to hospital admission or prolonging hospital stay. Mortality will be analyzed as total mortality as well as procedure-related mortality. The total costs of treatment will be evaluated by counting volumes and calculating unit prices. Including 120 patients on a 1:1 basis will have 80% power to detect an effect size of 0.5 on the EORTC QLQ-C30 global health scale, using a two group t-test with a 0.05 two-sided significance level. Differences in quality of life and morbidity will be analyzed using mixed-models repeated measures

  14. Reduction of cardiac and coronary artery doses in irradiation of left-sided breast cancer during inspiration breath hold. A planning study

    Energy Technology Data Exchange (ETDEWEB)

    Schoenecker, S.; Heinz, C.; Soehn, M.; Haimerl, W.; Corradini, S.; Pazos, M.; Belka, C.; Scheithauer, H. [University Hospital of Ludwig-Maximilian-University, Department of Radiation Oncology, Munich (Germany)

    2016-11-15

    The radiation dose received by the heart during adjuvant left-sided breast irradiation plays a crucial role in development of late toxicity. Although the absolute risk of cardiotoxicity can be reduced with modern irradiation techniques, cardiotoxic chemotherapy increases the risk of late damage. Thus, the radiation dose to the heart should be minimized. This study evaluated the influence of different amplitudes of inspiration breath hold (IBH) during simulated left-sided breast irradiation on cardiac doses compared to free breathing (FB). CT data of 11 lung cancer patients were retrospectively used as left-sided pseudo-breast cancer cases. Two CT scans were used, one during IBH and one during FB, and two treatment plans were generated. Relevant heart, lung, and left anterior descending artery (LAD) parameters were derived from dose-volume histograms. The normal tissue complication probabilities (NTCPs) for the heart were calculated based on the relative seriality model. Inspiration depth was quantified using chest volume and diameter, and correlated thereafter to a possible sparing of heart tissue. Mean reduction of heart dose for IBH compared to FB was 40 % (1.65 vs. 0.99 Gy; p = 0.007). Maximum dose to the heart and LAD could be decreased by 33 % (p = 0.011) and 43 % (p = 0.024), respectively. The mean anteroposterior shift was 5 mm (range 0.9-9.5 mm). Significant negative correlations between the relative change in LAD mean dose and the mean thoracic diameter and volume change, as well as with the absolute change in thoracic diameter were seen. The NTCP for cardiac mortality could be decreased by about 78 % (p = 0.017). For left-sided breast cancer patients, cardiac doses can be significantly decreased with tangential irradiation and IBH. (orig.) [German] Die Herzdosisbelastung spielt eine entscheidende Rolle in der Entwicklung einer kardialen Spaettoxizitaet nach der adjuvanten Strahlentherapie von linksseitigem Brustkrebs. Obwohl moderne strahlentherapeutische

  15. Comparison of conventional inserts and an add-on electron MLC for chest wall irradiation of left-sided breast cancer

    International Nuclear Information System (INIS)

    Vatanen, Tero; Lahtinen, Tapani; Traneus, Erik

    2009-01-01

    Background. Collimation of irregularly shaped clinical electron beams is currently based on electron inserts made of low melting point alloys. The present investigation compares a conventional electron applicator with insert and add-on eMLC-based dose distributions in the postoperative chest wall irradiation of left-sided breast cancer. Material and methods. Voxel Monte Carlo++ (VMC++) calculated dose distributions related to electron fields were compared with 10 left-sided breast cancer patients after radical mastectomy. The prescription dose was 50 Gy at a build-up maximum. The same dose was prescribed for the ipsilateral axillary, parasternal and supraclavicular lymph nodes that were treated with photons and calculated with a pencil beam algorithm. The insert beams were shaped with 1.5 cm thick Wood's metal electron inserts in an electron applicator of a Varian 2100 C/D linac. Doses for the eMLC-shaped beams were calculated for an eMLC prototype with 2 cm thick and 5 mm wide steel leaves. The same collimator-to-surface distance (CSD) of 5.8 cm was used for both collimators. Results. The mean PTV dose was slightly higher for the eMLC plans (50.7 vs 49.5 Gy, p<0.001, respectively). The maximum doses assessed by D5% for the eMLC and insert were 60.9 and 59.1 Gy (p<0.001). The difference was due to the slightly higher doses near the field edges for the eMLC. The left lung V20 volumes were 34.5% and 34.0% (p<0.001). There was only a marginal difference in heart doses. Discussion: Despite a slight increase of maximum dose in PTV the add-on electron MLC for chest wall irradiation results in practically no differences in dose distributions compared with the present insert-based collimation

  16. Cardiac dose reduction with deep inspiration breath hold for left-sided breast cancer radiotherapy patients with and without regional nodal irradiation.

    Science.gov (United States)

    Yeung, Rosanna; Conroy, Leigh; Long, Karen; Walrath, Daphne; Li, Haocheng; Smith, Wendy; Hudson, Alana; Phan, Tien

    2015-09-22

    Deep inspiration breath hold (DIBH) reduces heart and left anterior descending artery (LAD) dose during left-sided breast radiation therapy (RT); however there is limited information about which patients derive the most benefit from DIBH. The primary objective of this study was to determine which patients benefit the most from DIBH by comparing percent reduction in mean cardiac dose conferred by DIBH for patients treated with whole breast RT ± boost (WBRT) versus those receiving breast/chest wall plus regional nodal irradiation, including internal mammary chain (IMC) nodes (B/CWRT + RNI) using a modified wide tangent technique. A secondary objective was to determine if DIBH was required to meet a proposed heart dose constraint of Dmean irradiation.

  17. W/kit gene required for interstitial cells of Cajal and for intestinal pacemaker activity

    DEFF Research Database (Denmark)

    Huizinga, J D; Thuneberg, L; Klüppel, M

    1995-01-01

    The pacemaker activity in the mammalian gut is responsible for generating anally propagating phasic contractions. The cellular basis for this intrinsic activity is unknown. The smooth muscle cells of the external muscle layers and the innervated cellular network of interstitial cells of Cajal......, which is closely associated with the external muscle layers of the mammalian gut, have both been proposed to stimulate pacemaker activity. The interstitial cells of Cajal were identified in the last century but their developmental origin and function have remained unclear. Here we show...... of Cajal associated with Auerbach's nerve plexus and intestinal pacemaker activity....

  18. A novel method of placing right ventricular leads in patients with persistent left superior vena cava using a conventional j stylet.

    Science.gov (United States)

    Mora, Guillermo

    2014-03-01

    Locating pacemaker electrodes can become complicated by congenital abnormalities such as persistent left superior vena cava (LSVC). To evaluate a technique for the implanting of ventricular electrode in patients with persistent LSVC. The study was carried out from June 2001 to June 2010 involving all patients who were admitted to the Hospital Universitario Mayor, Instituto de Corazon de Bogota and Hospital Universitario Clinica San Rafael (Bogota-Colombia) for implanting pacemakers or cardiac defibrillators. LSVC was diagnosed by fluoroscopic observation (anterior-posterior view) of the course of the stylet. Four steps were followed: 1) Move the electrode with a straight stylet to the right atrium. 2) Change the straight stylet by a conventional J stylet and push the electrode to the lateral or anterolateral wall of the right atrium. 3) Remove the guide 3-5 cm and 4) Push the electrode which crosses the tricuspid valve into the right ventricle and finally deploy the active fixation mechanism. A total of 1198 patients were admitted for pacemaker or cardiac defibrillator implant during the 9-year study period, 1114 received a left subclavian venous approach. There were 573 males and 541 females. Persistent LSVC was found in five patients (0.45%) Fluoroscopy time for implanting the ventricular electrode ranged from 60 to 250 seconds, 40 to 92 minutes being taken to complete the whole procedure. We present a simple and rapid technique for electrode placement in patients with LSVC using usual J guide and active fixation electrodes with high success.

  19. [A case of traumatic middle meningeal arteriovenous fistula on the side of the head opposite to the injured side].

    Science.gov (United States)

    Takeuchi, Satoru; Takasato, Yoshio; Masaoka, Hiroyuki; Hayakawa, Takanori; Otani, Naoki; Yoshino, Yoshikazu; Yatsushige, Hiroshi; Sugawara, Takashi; Aoyagi, Chikashi; Suzuki, Go

    2009-10-01

    A rare case of a traumatic middle meningeal arteriovenous fistula on the side of the head opposite to the injured side was reported. A 21-year-old man was admitted to our hospital after a traffic accident in which the right side of his head was hit. CT scans and MR images on admission showed a right temporal bone fracture, traumatic subarachnoid hemorrhage, and a left frontal lobe contusion. Three months after the head injury, he complained of tinnitus and exophthalmos. One year after the head injury, left external carotid angiograms showed a dural arteriovenous fistula fed by the left dilated middle meningeal artery and draining into the middle meningeal vein. Early filling of the sphenoparietal sinus, cavernous sinus, superior ophthalmic vein, and the cortical vein were also detected. Transarterial embolization of the left middle meningeal fistula was performed, resulting in the disappearance of the lesion. The postoperative course was uneventful.

  20. Prevention of disease progression by cardiac resynchronization therapy in patients with asymptomatic or mildly symptomatic left ventricular dysfunction: insights from the European cohort of the REVERSE (Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction) trial

    DEFF Research Database (Denmark)

    Daubert, Claude; Gold, Michael R; Abraham, William T

    2009-01-01

    were decreased in this patient population in New York Heart Association functional classes I or II. These observations suggest that CRT prevents the progression of disease in patients with asymptomatic or mildly symptomatic LV dysfunction. (REsynchronization reVErses Remodeling in Systolic Left v......OBJECTIVES: The aim of this study was to determine the long-term effects of cardiac resynchronization therapy (CRT) in the European cohort of patients enrolled in the REVERSE (Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction) trial. BACKGROUND: Previous data suggest...... that CRT slows disease progression and improves the outcomes of asymptomatic or mildly symptomatic patients with left ventricular (LV) dysfunction and a wide QRS complex. METHODS: We randomly assigned 262 recipients of CRT pacemakers or defibrillators, with QRS > or =120 ms and LV ejection fraction...

  1. Quality assessment of pacemaker implantations in Denmark

    DEFF Research Database (Denmark)

    Møller, M; Arnsbo, P; Asklund, Mogens

    2002-01-01

    AIMS: Quality assessment of therapeutic procedures is essential to insure a cost-effective health care system. Pacemaker implantation is a common procedure with more than 500,000 implantations world-wide per year, but the general complication rate is not well described. We studied procedure relat...

  2. Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 Tesla?

    Science.gov (United States)

    Vahlhaus, C; Sommer, T; Lewalter, T; Schimpf, R; Schumacher, B; Jung, W; Lüderitz, B

    2001-04-01

    The safety and feasibility of magnetic resonance imaging (MRI) in patients with cardiac pacemakers is an issue of gaining significance. The effect of MRI on patients' pacemaker systems has only been analyzed retrospectively in some case reports. Therefore, this study prospectively investigated if MRI causes irreversible changes in patients' pacemaker systems. The effect of MRI at 0.5 Tesla on sensing and stimulation thresholds, lead impedance and battery voltage, current, and impedance was estimated during 34 MRI examinations in 32 patients with implanted pacemakers. After measurements at baseline and with documentation of intrinsic rhythm and modification of the pacing mode, patients underwent MRI. The rest of the function time of the pacemaker was calculated. Measurements were again performed after 99.5 +/- 29.6 minutes (mean +/- SD), immediately after MRI examination, and 3 months later. Lead impedance and sensing and stimulation thresholds did not change after MRI. Battery voltage decreased immediately after MRI and recovered 3 months later. Battery current and impedance tended to increase. The calculated rest of function time did not change immediately after MRI. MRI affected neither pacemaker programmed data, nor the ability to interrogate, program, or use telemetry. Surprisingly, in the gantry of the scanner, temporary deactivation of the reed switch occurred in 12 of 32 patients when positioned in the center of the magnetic field. Missing activation of the reed switch through the static magnetic field at 0.5 Tesla is not unusual. MRI at 0.5 Tesla does not cause irreversible changes in patients' pacemaker systems.

  3. Interference of apex locator, pulp tester and diathermy on pacemaker function

    OpenAIRE

    Narayanan Sriman; V Prabhakar; J S Bhuvaneswaran; N Subha

    2015-01-01

    Aim: The purpose of this study was to evaluate the effects of three electronic apex locators (EAL), electric pulp tester (EPT) and diathermy on pacemaker function in vitro. Materials and Methods: Three EALs: Root ZX (J. Morita Co., Tustin, CA, U.S.A.), Propex (Dentsply), Mini Apex locator (SybronEndo, Anaheim, CA, USA), EPT (Parkell pulp vitality tester Farmingdale, NY, USA) and Diathermy (Neomed 250 B) were tested for any interference with one pacemaker (A medtronic kappa KVDD901-serial ...

  4. Exploring the mechanism of action of the sperm-triggered calcium-wave pacemaker in ascidian zygotes.

    Science.gov (United States)

    Carroll, Michael; Levasseur, Mark; Wood, Chris; Whitaker, Michael; Jones, Keith T; McDougall, Alex

    2003-12-15

    In ascidians, as in mammals, sperm trigger repetitive Ca2+-waves that originate from cortical pacemakers situated in the vegetal hemisphere of the zygotes. In ascidians, a vegetal protrusion termed the contraction pole (CP) acts as the Ca2+-wave pacemaker, but the mechanism that underlies the generation of a Ca2+-wave pacemaker is not known. Here, we tested four hypotheses to determine which factors at the CP are involved in setting the pace of the ascidian Ca2+-wave pacemaker: (1) localized Ca2+ influx; (2) accumulation of phosphatidylinositol (4,5)bisphosphate [PtdIns(4,5)P2]; (3) accumulation of cortical endoplasmic reticulum (cER); and (4) enrichment of the sperm activating factor. We developed a method of dynamically monitoring the location of the CP during fertilization using a plekstrin homology (PH) domain from phospholipase Cdelta1 coupled to green fluorescent protein (GFP) that binds PtdIns(4,5)P2. We found that eggs in Ca2+-free sea water displayed Ca2+ waves that originated from the CP, showing that enhanced CP Ca2+ influx does not determine the origin of the pacemaker. Also, disruption of the PH::GFP-labelled CP once it had formed did not dislodge the Ca2+-wave pacemaker from that site. Next, when we prevented the accumulation of cER at the CP, all of the Ca2+ waves came from the site of sperm-egg fusion and the frequency of Ca2+ oscillations was unaltered. These data show that local Ca2+ influx, the accumulation of PtdIns(4,5)P2 and cER at the CP are not required for Ca2+-wave pacemaker function and instead suggest that a factor associated with the sperm determines the site of the Ca2+-wave pacemaker. Finally, when we injected ascidian sperm extract into the centre of unfertilized ascidian eggs that had been treated with microfilament- and microtubule-disrupting drugs, all the Ca2+ waves still originated from near the plasma membrane, showing that the sperm factor does not require an intact cortex if it is enriched near the plasma membrane (PM). We

  5. Identification and functional characterization of cardiac pacemaker cells in zebrafish.

    Directory of Open Access Journals (Sweden)

    Federico Tessadori

    Full Text Available In the mammalian heart a conduction system of nodes and conducting cells generates and transduces the electrical signals evoking myocardial contractions. Specialized pacemaker cells initiating and controlling cardiac contraction rhythmicity are localized in an anatomically identifiable structure of myocardial origin, the sinus node. We previously showed that in mammalian embryos sinus node cells originate from cardiac progenitors expressing the transcription factors T-box transcription factor 3 (Tbx3 and Islet-1 (Isl1. Although cardiac development and function are strikingly conserved amongst animal classes, in lower vertebrates neither structural nor molecular distinguishable components of a conduction system have been identified, questioning its evolutionary origin. Here we show that zebrafish embryos lacking the LIM/homeodomain-containing transcription factor Isl1 display heart rate defects related to pacemaker dysfunction. Moreover, 3D reconstructions of gene expression patterns in the embryonic and adult zebrafish heart led us to uncover a previously unidentified, Isl1-positive and Tbx2b-positive region in the myocardium at the junction of the sinus venosus and atrium. Through their long interconnecting cellular protrusions the identified Isl1-positive cells form a ring-shaped structure. In vivo labeling of the Isl1-positive cells by transgenic technology allowed their isolation and electrophysiological characterization, revealing their unique pacemaker activity. In conclusion we demonstrate that Isl1-expressing cells, organized as a ring-shaped structure around the venous pole, hold the pacemaker function in the adult zebrafish heart. We have thereby identified an evolutionary conserved, structural and molecular distinguishable component of the cardiac conduction system in a lower vertebrate.

  6. Twenty-Seven Years Experience With Transvenous Pacemaker Implantation in Children Weighing <10 kg.

    Science.gov (United States)

    Konta, Laura; Chubb, Mark Henry; Bostock, Julian; Rogers, Jan; Rosenthal, Eric

    2016-02-01

    Epicardial pacemaker implantation is the favored approach in children weighing pacemaker implantation in neonates and infants from 1987. To date there have been no long-term follow-up reports of what is for many a controversial strategy. Between 1987 and 2003, 37 neonates and infants-median age 6.7 months (1 day to 3 years) and median weight 4.6 kg (2.7-10 kg)-had a permanent transvenous pacing system implanted. Pacing leads were placed into the right ventricular apex/outflow tract through a subclavian vein puncture with a redundant loop in the atrium. Three patients were lost to follow-up, 4 patients died from complications of cardiac surgery, and 2 patients had their system removed. At long-term follow-up in 28 patients at a median of 17.2 (range, 11.2-27.4) years, 10 patients have a single chamber ventricular pacemaker, 14 a dual chamber pacemaker, 3 a biventricular pacemaker, and 1 has a single chamber implantable cardioverter defibrillator. Subclavian vein patency was assessed in 26 patients. The overall subclavian vein occlusion rate was 10 of 13 (77%) 5 kg during long-term follow-up. After a median of 14.3 (range, 13.4-17.6) years of pacing, 7 patients continue with their original lead. Transvenous pacing in infants <10 kg results in encouraging short- and long-term clinical outcomes. Subclavian vein occlusion remains an important complication, occurring predominantly in those weighing <5 kg. © 2016 American Heart Association, Inc.

  7. 'Right-Sided' May-Thurner Syndrome

    International Nuclear Information System (INIS)

    Abboud, Georges; Midulla, Marco; Lions, Christophe; El Ngheoui, Ziad; Gengler, Laurent; Martinelli, Thomas; Beregi, Jean-Paul

    2010-01-01

    The May-Thurner syndrome is a well-known anatomical anomaly where the left common iliac vein (LCIV) is compressed between the right common iliac artery and the fifth vertebral body. This report describes the case of a 'right-sided' May-Thurner syndrome where the right common iliac vein (RCIV) is compressed by the left common iliac artery in a patient with a left-sided inferior vena cava (IVC). A 26-year-old woman was admitted to our institution with acute edema of the right lower limb. The diagnosis of May-Thurner syndrome was done by CT scan and confirmed by phlebography. An endovascular treatment with stenting was carried out, with good patency and clinical result at 12-month follow-up.

  8. Percutaneously injectable fetal pacemaker: electrodes, mechanical design and implantation.

    Science.gov (United States)

    Zhou, Li; Chmait, Ramen; Bar-Cohen, Yaniv; Peck, Raymond A; Loeb, Gerald E

    2012-01-01

    We are developing a self-contained cardiac pacemaker with a small, cylindrical shape (~3 × 20 mm) that permits it to be implanted percutaneously into a fetus to treat complete heart block and consequent hydrops fetalis, which is otherwise fatal. The device uses off-the-shelf components including a rechargeable lithium cell and a highly efficient relaxation oscillator encapsulated in epoxy and glass. A corkscrew electrode made from activated iridium can be screwed into the myocardium, followed by release of the pacemaker and a short, flexible lead entirely within the chest of the fetus to avoid dislodgement from fetal movement. The feasibility of implanting the device percutaneously under ultrasonic imaging guidance was demonstrated in acute adult rabbit experiments.

  9. Use of an active fixation lead and a subpectoral pacemaker pocket may not avoid Twiddler′s syndrome

    Directory of Open Access Journals (Sweden)

    Floris E A Udink ten Cate

    2012-01-01

    Full Text Available Manipulation of a pacemaker with consequent malfunction of the device has been called Twiddler′s syndrome. Use of active-fixation leads and subpectoral pacemaker pockets has been considered to help in avoiding this problem. We describe a child in whom twiddling was not prevented despite implantation of a lumenless atrial lead and insertion of the pacemaker generator in a subpectoral pocket.

  10. Positive pressure ventilation in a patient with a right upper lobar bronchocutaneous fistula: right upper bronchus occlusion using the cuff of a left-sided double lumen endobronchial tube.

    Science.gov (United States)

    Omori, Chieko; Toyama, Hiroaki; Takei, Yusuke; Ejima, Yutaka; Yamauchi, Masanori

    2017-08-01

    In patients with a bronchocutaneous fistula, positive pressure ventilation leads to air leakage and potential hypoxemia. A male patient with a right upper bronchocutaneous fistula was scheduled for esophageal reconstruction. His preoperative chest computed tomography image revealed aeration in the right middle and lower lobe, a large bulla in the left upper lobe, and pleural effusion and pneumonia in the left lower lobe. Therefore, left one-lung ventilation was considered to result in hypoxemia. Before anesthesia induction, the bronchocutaneous fistula was covered with gauze and film to prevent air leakage. After anesthesia induction, mask ventilation was performed with a peak positive pressure of 10 cmH 2 O. A left-sided double lumen endobronchial tube (DLT) was then inserted into the right main bronchus for occluding only the right superior bronchus, and two-lung ventilation was performed to minimize airway pressure and maintain oxygenation, which did not cause air leakage through the fistula. During anesthesia, no ventilation-related difficulty was faced. The method of inserting a left-sided DLT into the right main bronchus and occluding the right upper bronchus selectively by bronchial cuff is considered to be an option for mechanical ventilation in patients with a right upper bronchial fistula, as demonstrated in the present case.

  11. Keeping the Rhythm : Cardiac Pacemaker Cell Development

    NARCIS (Netherlands)

    Burkhard, S.B.

    2017-01-01

    The heart is the first organ to form and function in the developing vertebrate embryo. Its proper morphogenesis and function is crucial for survival. Here we focus on the development and characterization of a highly specialized subset of cardiac cells, the pacemaker cells. In the mammalian heart,

  12. Neutron absorbed dose in a pacemaker CMOS

    International Nuclear Information System (INIS)

    Borja H, C. G.; Guzman G, K. A.; Valero L, C.; Banuelos F, A.; Hernandez D, V. M.; Vega C, H. R.; Paredes G, L.

    2012-01-01

    The neutron spectrum and the absorbed dose in a Complementary Metal Oxide Semiconductor (CMOS), has been estimated using Monte Carlo methods. Eventually a person with a pacemaker becomes an oncology patient that must be treated in a linear accelerator. Pacemaker has integrated circuits as CMOS that are sensitive to intense and pulsed radiation fields. Above 7 MV therapeutic beam is contaminated with photoneutrons that could damage the CMOS. Here, the neutron spectrum and the absorbed dose in a CMOS cell was calculated, also the spectra were calculated in two point-like detectors in the room. Neutron spectrum in the CMOS cell shows a small peak between 0.1 to 1 MeV and a larger peak in the thermal region, joined by epithermal neutrons, same features were observed in the point-like detectors. The absorbed dose in the CMOS was 1.522 x 10 -17 Gy per neutron emitted by the source. (Author)

  13. Neutron absorbed dose in a pacemaker CMOS

    Energy Technology Data Exchange (ETDEWEB)

    Borja H, C. G.; Guzman G, K. A.; Valero L, C.; Banuelos F, A.; Hernandez D, V. M.; Vega C, H. R. [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico); Paredes G, L., E-mail: fermineutron@yahoo.com [ININ, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)

    2012-06-15

    The neutron spectrum and the absorbed dose in a Complementary Metal Oxide Semiconductor (CMOS), has been estimated using Monte Carlo methods. Eventually a person with a pacemaker becomes an oncology patient that must be treated in a linear accelerator. Pacemaker has integrated circuits as CMOS that are sensitive to intense and pulsed radiation fields. Above 7 MV therapeutic beam is contaminated with photoneutrons that could damage the CMOS. Here, the neutron spectrum and the absorbed dose in a CMOS cell was calculated, also the spectra were calculated in two point-like detectors in the room. Neutron spectrum in the CMOS cell shows a small peak between 0.1 to 1 MeV and a larger peak in the thermal region, joined by epithermal neutrons, same features were observed in the point-like detectors. The absorbed dose in the CMOS was 1.522 x 10{sup -17} Gy per neutron emitted by the source. (Author)

  14. Permanent and temporary pacemaker implantation after orthotopic heart transplantation

    Directory of Open Access Journals (Sweden)

    Bacal Fernando

    2000-01-01

    Full Text Available PURPOSE:To determine the indication for and incidence and evolution of temporary and permanent pacemaker implantation in cardiac transplant recipients. METHODS: A retrospective review of 114 patients who underwent orthotopic heart transplantation InCor (Heart Institute USP BR between March 1985 and May 1993. We studied the incidence of and indication for temporary pacing, the relationship between pacing and rejection, the need for pemanent pacing and the clinical follow-up. RESULTS: Fourteen of 114 (12%heart transplant recipients required temporary pacing and 4 of 114 (3.5% patients required permanent pacing. The indication for temporary pacing was sinus node dysfunction in 11 patients (78.5% and atrioventricular (AV block in 3 patients (21.4%. The indication for permanent pacemaker implantation was sinus node dysfunction in 3 patients (75% and atrioventricular (AV block in 1 patient (25%. We observed rejection in 3 patients (21.4% who required temporary pacing and in 2 patients (50% who required permanent pacing. The previous use of amiodarone was observed in 10 patients (71.4% with temporary pacing. Seven of the 14 patients (50% died during follow-up. CONCLUSION: Sinus node dysfunction was the principal indication for temporary and permanent pacemaker implantation in cardiac transplant recipients. The need for pacing was related to worse prognosis after cardiac transplantation.

  15. Modeling and experimental verification of a fan-folded vibration energy harvester for leadless pacemakers

    Science.gov (United States)

    Ansari, M. H.; Karami, M. Amin

    2016-03-01

    This paper studies energy harvesting from heartbeat vibrations for powering leadless pacemakers. Unlike traditional pacemakers, leadless pacemakers are implanted inside the heart and the pacemaker is in direct contact with the myocardium. A leadless pacemaker is in the shape of a cylinder. Thus, in order to utilize the available 3-dimensional space for the energy harvester, we choose a fan-folded 3D energy harvester. The proposed device consists of several piezoelectric beams stacked on top of each other. The volume of the energy harvester is 1 cm3 and its dimensions are 2 cm × 0.5 cm × 1 cm. Although high natural frequency is generally a major concern with micro-scale energy harvesters, by utilizing the fan-folded geometry and adding tip mass and link mass to the configuration, we reduced the natural frequency to the desired range. This fan-folded design makes it possible to generate more than 10 μ W of power per cubic centimeter. The proposed device is compatible with Magnetic Resonance Imaging. Although the proposed device is a linear energy harvester, it is relatively insensitive to the heart rate. The natural frequencies and the mode shapes of the device are calculated analytically. The accuracy of the analytical model is verified by experimental investigations. We use a closed loop shaker system to precisely replicate heartbeat vibrations in vitro.

  16. Laterotopic representation of left-right information onto the dorso-ventral axis of a zebrafish midbrain target nucleus.

    Science.gov (United States)

    Aizawa, Hidenori; Bianco, Isaac H; Hamaoka, Takanori; Miyashita, Toshio; Uemura, Osamu; Concha, Miguel L; Russell, Claire; Wilson, Stephen W; Okamoto, Hitoshi

    2005-02-08

    The habenulae are part of an evolutionarily highly conserved limbic-system conduction pathway that connects telencephalic nuclei to the interpeduncular nucleus (IPN) of the midbrain . In zebrafish, unilateral activation of the Nodal signaling pathway in the left brain specifies the laterality of the asymmetry of habenular size . We show "laterotopy" in the habenulo-interpeduncular projection in zebrafish, i.e., the stereotypic, topographic projection of left-sided habenular axons to the dorsal region of the IPN and of right-sided habenular axons to the ventral IPN. This asymmetric projection is accounted for by a prominent left-right (LR) difference in the size ratio of the medial and lateral habenular sub-nuclei, each of which specifically projects either to ventral or dorsal IPN targets. Asymmetric Nodal signaling directs the orientation of laterotopy but is dispensable for the establishment of laterotopy itself. Our results reveal a mechanism by which information distributed between left and right sides of the brain can be transmitted bilaterally without loss of LR coding, which may play a crucial role in functional lateralization of the vertebrate brain .

  17. Left-Right Asymmetric Morphogenesis in the Xenopus Digestive System

    Science.gov (United States)

    Muller, Jennifer K.; Prather, D.R.; Nascone-Yoder, N. M.

    2003-01-01

    The morphogenetic mechanisms by which developing organs become left-right asymmetric entities are unknown. To investigate this issue, we compared the roles of the left and right sides of the Xenopus embryo during the development of anatomic asymmetries in the digestive system. Although both sides contribute equivalently to each of the individual digestive organs, during the initial looping of the primitive gut tube, the left side assumes concave topologies where the right side becomes convex. Of interest, the concave surfaces of the gut tube correlate with expression of the LR gene, Pitx2, and ectopic Pitx2 mRNA induces ectopic concavities in a localized manner. A morphometric comparison of the prospective concave and convex surfaces of the gut tube reveals striking disparities in their rate of elongation but no significant differences in cell proliferation. These results provide insight into the nature of symmetry-breaking morphogenetic events during left-right asymmetric organ development. ?? 2003 Wiley-Liss, Inc.

  18. Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators.

    Science.gov (United States)

    Wolber, Thomas; Ryf, Salome; Binggeli, Christian; Holzmeister, Johannes; Brunckhorst, Corinna; Luechinger, Roger; Duru, Firat

    2007-01-01

    Magnetic fields may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets, which are small in size but produce strong magnetic fields, have become widely available in recent years. Therefore, NdFeB magnets may be associated with an emerging risk of device interference. We conducted a clinical study to evaluate the potential of small NdFeB magnets to interfere with cardiac pacemakers and ICDs. The effect of four NdFeB magnets (two spherical magnets 8 and 10 mm in diameter, a necklace made of 45 spherical magnets, and a magnetic name tag) was tested in forty-one ambulatory patients with a pacemaker and 29 patients with an ICD. The maximum distance at which the magnetic switch of a device was influenced was observed. Magnetic interference was observed in all patients. The maximum distance resulting in device interference was 3 cm. No significant differences were found with respect to device manufacturer and device types. Small NdFeB magnets may cause interference with cardiac pacemakers and ICDs. Patients should be cautioned about the interference risk associated with NdFeB magnets during daily life.

  19. Clinical features and changes in epidemiology of infective endocarditis on pacemaker devices over a 27-year period (1987-2013).

    Science.gov (United States)

    Carrasco, Francisco; Anguita, Manuel; Ruiz, Martín; Castillo, Juan Carlos; Delgado, Mónica; Mesa, Dolores; Romo, Elias; Pan, Manuel; Suárez de Lezo, Jose

    2016-06-01

    Use of cardiac pacing devices has grown in recent years. Our aim was to evaluate changes in epidemiology and clinical features of infective endocarditis (IE) involving pacemaker devices in a large series of IE over the last 27 years (1987-2013). From 1987 to December 2013, 413 consecutive IE cases were diagnosed in our hospital. During this period, 7424 pacemaker devices were implanted (6917 pacemakers, 239 implantable cardiac defibrillators, 158 resynchronization devices, and 110 resynchronization/defibrillator devices). All consecutive cases of IE on pacemaker devices were included and analysed. Infective endocarditis on pacemaker devices represented 6.1% of all endocarditis cases (25 patients), affecting 3.6/1000 of all implanted pacemakers. Its proportion increased from 1.25% of all endocarditis in 1987-1993 to 4.08% in 1994-2000, 7.69% in 2001-2007 and 9.32% in 2008-2013 (P pacemaker implants in the period of 1987-1993 to 2.5/1000 in 1994-2000, 3.3/1000 in 2001-2007 and 4.5/1000 implanted devices in 2008-2013 (P pacemaker devices has shown an increasing incidence during the past decades, representing almost 10% of all IE in the last 6 years. This is a severe disease, with a high rate of severe complications and requiring removal of device in most cases. In spite of therapy, early mortality is high. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2015. For permissions please email: journals.permissions@oup.com.

  20. Low frequency magnetic emissions and resulting induced voltages in a pacemaker by iPod portable music players

    Directory of Open Access Journals (Sweden)

    Bassen Howard

    2008-02-01

    Full Text Available Abstract Background Recently, malfunctioning of a cardiac pacemaker electromagnetic, caused by electromagnetic interference (EMI by fields emitted by personal portable music players was highly publicized around the world. A clinical study of one patient was performed and two types of interference were observed when the clinicians placed a pacemaker programming head and an iPod were placed adjacent to the patient's implanted pacemaker. The authors concluded that "Warning labels may be needed to avoid close contact between pacemakers and iPods". We performed an in-vitro study to evaluate these claims of EMI and present our findings of no-effects" in this paper. Methods We performed in-vitro evaluations of the low frequency magnetic field emissions from various models of the Apple Inc. iPod music player. We measured magnetic field emissions with a 3-coil sensor (diameter of 3.5 cm placed within 1 cm of the surface of the player. Highly localized fields were observed (only existing in a one square cm area. We also measured the voltages induced inside an 'instrumented-can' pacemaker with two standard unipolar leads. Each iPod was placed in the air, 2.7 cm above the pacemaker case. The pacemaker case and leads were placed in a saline filled torso simulator per pacemaker electromagnetic compatibility standard ANSI/AAMI PC69:2000. Voltages inside the can were measured. Results Emissions were strongest (≈ 0.2 μT pp near a few localized points on the cases of the two iPods with hard drives. Emissions consisted of 100 kHz sinusoidal signal with lower frequency (20 msec wide pulsed amplitude modulation. Voltages induced in the iPods were below the noise level of our instruments (0.5 mV pp in the 0 – 1 kHz band or 2 mV pp in the 0 – 5 MHz bandwidth. Conclusion Our measurements of the magnitude and the spatial distribution of low frequency magnetic flux density emissions by 4 different models of iPod portable music players. Levels of less than 0.2