  1. Fast and accurate: high-speed metrological large-range AFM for surface and nanometrology

    Science.gov (United States)

    Dai, Gaoliang; Koenders, Ludger; Fluegge, Jens; Hemmleb, Matthias

    2018-05-01

    Low measurement speed remains a major shortcoming of the scanning probe microscopic technique. It not only leads to a low measurement throughput, but a significant measurement drift over the long measurement time needed (up to hours or even days). To overcome this challenge, PTB, the national metrology institute of Germany, has developed a high-speed metrological large-range atomic force microscope (HS Met. LR-AFM) capable of measuring speeds up to 1 mm s‑1. This paper has introduced the design concept in detail. After modelling scanning probe microscopic measurements, our results suggest that the signal spectrum of the surface to be measured is the spatial spectrum of the surface scaled by the scanning speed. The higher the scanning speed , the broader the spectrum to be measured. To realise an accurate HS Met. LR-AFM, our solution is to combine different stages/sensors synchronously in measurements, which provide a much larger spectrum area for high-speed measurement capability. Two application examples have been demonstrated. The first is a new concept called reference areal surface metrology. Using the developed HS Met. LR-AFM, surfaces are measured accurately and traceably at a speed of 500 µm s‑1 and the results are applied as a reference 3D data map of the surfaces. By correlating the reference 3D data sets and 3D data sets of tools under calibration, which are measured at the same surface, it has the potential to comprehensively characterise the tools, for instance, the spectrum properties of the tools. The investigation results of two commercial confocal microscopes are demonstrated, indicating very promising results. The second example is the calibration of a kind of 3D nano standard, which has spatially distributed landmarks, i.e. special unique features defined by 3D-coordinates. Experimental investigations confirmed that the calibration accuracy is maintained at a measurement speed of 100 µm s‑1, which improves the calibration efficiency by a

  2. Design and control of multi-actuated atomic force microscope for large-range and high-speed imaging

    Energy Technology Data Exchange (ETDEWEB)

    Soltani Bozchalooi, I.; Careaga Houck, A. [Department of Mechanical Engineering, Massachusetts Institute of Technology, 77 Massachusetts Avenue, Cambridge, MA 02139 (United States); AlGhamdi, J. [Department of Mechanical Engineering, Massachusetts Institute of Technology, 77 Massachusetts Avenue, Cambridge, MA 02139 (United States); Department of Chemistry, College of Science, University of Dammam, Dammam (Saudi Arabia); Youcef-Toumi, K. [Department of Mechanical Engineering, Massachusetts Institute of Technology, 77 Massachusetts Avenue, Cambridge, MA 02139 (United States)

    2016-01-15

    This paper presents the design and control of a high-speed and large-range atomic force microscopy (AFM). A multi-actuation scheme is proposed where several nano-positioners cooperate to achieve the range and speed requirements. A simple data-based control design methodology is presented to effectively operate the AFM scanner components. The proposed controllers compensate for the coupled dynamics and divide the positioning responsibilities between the scanner components. As a result, the multi-actuated scanner behavior is equivalent to that of a single X–Y–Z positioner with large range and high speed. The scanner of the designed AFM is composed of five nano-positioners, features 6 μm out-of-plane and 120 μm lateral ranges and is capable of high-speed operation. The presented AFM has a modular design with laser spot size of 3.5 μm suitable for small cantilever, an optical view of the sample and probe, a conveniently large waterproof sample stage and a 20 MHz data throughput for high resolution image acquisition at high imaging speeds. This AFM is used to visualize etching of calcite in a solution of sulfuric acid. Layer-by-layer dissolution and pit formation along the crystalline lines in a low pH environment is observed in real time. - Highlights: • High-speed AFM imaging is extended to large lateral and vertical scan ranges. • A general multi-actuation approach to atomic force microscopy is presented. • A high-speed AFM is designed and implemented based on the proposed method. • Multi-actuator control is designed auxiliary to a PID unit to maintain flexibility. • Influence of calcite crystal structure on dissolution is visualized in video form.

  3. Multi-actuation and PI control: A simple recipe for high-speed and large-range atomic force microscopy

    Energy Technology Data Exchange (ETDEWEB)

    Soltani Bozchalooi, I., E-mail: isoltani@mit.edu; Youcef-Toumi, K.

    2014-11-15

    High speed atomic force microscopy enables observation of dynamic nano-scale processes. However, maintaining a minimal interaction force between the sample and the probe is challenging at high speed specially when using conventional piezo-tubes. While rigid AFM scanners are operational at high speeds with the drawback of reduced tracking range, multi-actuation schemes have shown potential for high-speed and large-range imaging. Here we present a method to seamlessly incorporate additional actuators into conventional AFMs. The equivalent behavior of the resulting multi-actuated setup resembles that of a single high-speed and large-range actuator with maximally flat frequency response. To achieve this, the dynamics of the individual actuators and their couplings are treated through a simple control scheme. Upon the implementation of the proposed technique, commonly used PI controllers are able to meet the requirements of high-speed imaging. This forms an ideal platform for retroactive enhancement of existing AFMs with minimal cost and without compromise on the tracking range. A conventional AFM with tube scanner is retroactively enhanced through the proposed method and shows an order of magnitude improvement in closed loop bandwidth performance while maintaining large range. The effectiveness of the method is demonstrated on various types of samples imaged in contact and tapping modes, in air and in liquid. - Highlights: • We present a novel method to incorporate extra actuators into conventional AFMs. • A maximally flat frequency response is achieved for the out of plane piezo-motion. • Commonly used PI or PID control is enabled to handle high speed AFM imaging. • An order of magnitude improvement in closed loop bandwidth performance is obtained. • High speed imaging is achieved on a large range piezo-tube.

  4. Multi-actuation and PI control: A simple recipe for high-speed and large-range atomic force microscopy

    International Nuclear Information System (INIS)

    Soltani Bozchalooi, I.; Youcef-Toumi, K.

    2014-01-01

    High speed atomic force microscopy enables observation of dynamic nano-scale processes. However, maintaining a minimal interaction force between the sample and the probe is challenging at high speed specially when using conventional piezo-tubes. While rigid AFM scanners are operational at high speeds with the drawback of reduced tracking range, multi-actuation schemes have shown potential for high-speed and large-range imaging. Here we present a method to seamlessly incorporate additional actuators into conventional AFMs. The equivalent behavior of the resulting multi-actuated setup resembles that of a single high-speed and large-range actuator with maximally flat frequency response. To achieve this, the dynamics of the individual actuators and their couplings are treated through a simple control scheme. Upon the implementation of the proposed technique, commonly used PI controllers are able to meet the requirements of high-speed imaging. This forms an ideal platform for retroactive enhancement of existing AFMs with minimal cost and without compromise on the tracking range. A conventional AFM with tube scanner is retroactively enhanced through the proposed method and shows an order of magnitude improvement in closed loop bandwidth performance while maintaining large range. The effectiveness of the method is demonstrated on various types of samples imaged in contact and tapping modes, in air and in liquid. - Highlights: • We present a novel method to incorporate extra actuators into conventional AFMs. • A maximally flat frequency response is achieved for the out of plane piezo-motion. • Commonly used PI or PID control is enabled to handle high speed AFM imaging. • An order of magnitude improvement in closed loop bandwidth performance is obtained. • High speed imaging is achieved on a large range piezo-tube

  5. Ultra-large scale AFM of lipid droplet arrays: investigating the ink transfer volume in dip pen nanolithography

    International Nuclear Information System (INIS)

    Förste, Alexander; Pfirrmann, Marco; Sachs, Johannes; Gröger, Roland; Walheim, Stefan; Brinkmann, Falko; Hirtz, Michael; Fuchs, Harald; Schimmel, Thomas

    2015-01-01

    There are only few quantitative studies commenting on the writing process in dip-pen nanolithography with lipids. Lipids are important carrier ink molecules for the delivery of bio-functional patters in bio-nanotechnology. In order to better understand and control the writing process, more information on the transfer of lipid material from the tip to the substrate is needed. The dependence of the transferred ink volume on the dwell time of the tip on the substrate was investigated by topography measurements with an atomic force microscope (AFM) that is characterized by an ultra-large scan range of 800 × 800 μm 2 . For this purpose arrays of dots of the phospholipid1,2-dioleoyl-sn-glycero-3-phosphocholine were written onto planar glass substrates and the resulting pattern was imaged by large scan area AFM. Two writing regimes were identified, characterized of either a steady decline or a constant ink volume transfer per dot feature. For the steady state ink transfer, a linear relationship between the dwell time and the dot volume was determined, which is characterized by a flow rate of about 16 femtoliters per second. A dependence of the ink transport from the length of pauses before and in between writing the structures was observed and should be taken into account during pattern design when aiming at best writing homogeneity. The ultra-large scan range of the utilized AFM allowed for a simultaneous study of the entire preparation area of almost 1 mm 2 , yielding good statistic results. (paper)

  6. Ultra-large scale AFM of lipid droplet arrays: investigating the ink transfer volume in dip pen nanolithography

    Science.gov (United States)

    Förste, Alexander; Pfirrmann, Marco; Sachs, Johannes; Gröger, Roland; Walheim, Stefan; Brinkmann, Falko; Hirtz, Michael; Fuchs, Harald; Schimmel, Thomas

    2015-05-01

    There are only few quantitative studies commenting on the writing process in dip-pen nanolithography with lipids. Lipids are important carrier ink molecules for the delivery of bio-functional patters in bio-nanotechnology. In order to better understand and control the writing process, more information on the transfer of lipid material from the tip to the substrate is needed. The dependence of the transferred ink volume on the dwell time of the tip on the substrate was investigated by topography measurements with an atomic force microscope (AFM) that is characterized by an ultra-large scan range of 800 × 800 μm2. For this purpose arrays of dots of the phospholipid1,2-dioleoyl-sn-glycero-3-phosphocholine were written onto planar glass substrates and the resulting pattern was imaged by large scan area AFM. Two writing regimes were identified, characterized of either a steady decline or a constant ink volume transfer per dot feature. For the steady state ink transfer, a linear relationship between the dwell time and the dot volume was determined, which is characterized by a flow rate of about 16 femtoliters per second. A dependence of the ink transport from the length of pauses before and in between writing the structures was observed and should be taken into account during pattern design when aiming at best writing homogeneity. The ultra-large scan range of the utilized AFM allowed for a simultaneous study of the entire preparation area of almost 1 mm2, yielding good statistic results.

  7. Evaluation of defect density by top-view large scale AFM on metamorphic structures grown by MOVPE

    Energy Technology Data Exchange (ETDEWEB)

    Gocalinska, Agnieszka, E-mail: agnieszka.gocalinska@tyndall.ie; Manganaro, Marina; Dimastrodonato, Valeria; Pelucchi, Emanuele

    2015-09-15

    Highlights: • Metamorphic buffer layers of In{sub x}Ga{sub 1−x}As were grown by MOVPE and characterised by AFM and TEM. • It was found that AFM provides sufficient information to estimate threading defect density in metamorphic structures, even when significant roughness is present. • When planar-view TEM is lacking, a combination of cross-sectional TEM and large scale AFM can provide good evaluation of the material quality. • It is fast, cheap and non-destructive – can be very useful in development process of complicated structures, requiring multiple test growths and characterisation. - Abstract: We demonstrate an atomic force microscopy based method for estimation of defect density by identification of threading dislocations on a non-flat surface resulting from metamorphic growth. The discussed technique can be applied as an everyday evaluation tool for the quality of epitaxial structures and allow for cost reduction, as it lessens the amount of the transmission electron microscopy analysis required at the early stages of projects. Metamorphic structures with low surface defectivities (below 10{sup 6}) were developed successfully with the application of the technique, proving its usefulness in process optimisation.

  8. High-field magnetoconductance in La-Sr manganites of FM and AFM ground states

    Science.gov (United States)

    Jirák, Zdeněk; Kaman, Ondřej; Knížek, Karel; Levinský, Petr; Míšek, Martin; Veverka, Pavel; Hejtmánek, Jiří

    2018-06-01

    Large-grain La1-xSrxMnO3 ceramic samples of compositions x = 0.45 and 0.55, representing the ferromagnetic (FM) and A-type antiferromagnetic (AFM) ground states, were produced via classical sintering at 1500 °C of cold-pressed sol-gel prepared single-phase nanoparticles. Using the same precursors, nanogranular forms of both manganite ceramics were prepared by fast spark plasma sintering at low temperature of 900 °C, which limits the growth of crystal grains. The magnetotransport of both the bulk and nanogranular forms was investigated in a broad range of magnetic fields up to 130 kOe and analyzed on the basis of detailed magnetic measurements. Both the large-grain and nanogranular systems with x = 0.45, possessing a pure FM state with similar Curie tempereature TC ≈ 345 K), show nearly the same conductivity enhancement in external fields when expressed relatively to the zero-field values. This positive magnetoconductance (MC) can be separated into two terms: (i) the hysteretic low-field MC that reflects the field-induced orientation of magnetic moments of individual grains, and (ii) the high-field MC that depends linearly on external field. In the case of large-grain ceramics with x = 0.55, a partially ordered FM state formed below TC = 264 K is replaced by pure A-type AFM ground state below 204 K. This A-type AFM state is characterized by positive magnetoconductance that is essentially of quadratic dependence on external field in the investigated range up to 130 kOe. On contrary, the nanogranular product with x = 0.55 exhibits a mixed FM/AFM state at low temperatures, and, as a consequence, its magnetotransport combines the features of FM and A-type AFM systems, in which the quadratic term is much enhanced and clearly dominates at high fields. For interpretation of observed behaviors, the theory of grain-boundary tunneling is revisited.

  9. Characterisation of tissue factor-bearing extracellular vesicles with AFM: comparison of air-tapping-mode AFM and liquid Peak Force AFM

    Directory of Open Access Journals (Sweden)

    Julie Hardij

    2013-08-01

    Full Text Available Introduction: Extracellular vesicles (EVs are shed from cells and carry markers of the parent cells. Vesicles derived from cancer cells reach the bloodstream and locally influence important physiological processes. It has been previously shown that procoagulant vesicles are circulating in patients’ fluids. These EVs are therefore considered as promising biomarkers for the thrombotic risk. Because of their small size, classical methods such as flow cytometry suffer from limitation for their characterisation. Atomic force microscopy (AFM has been proposed as a promising complementary method for the characterisation of EVs. Objectives: The objectives of this study are: (a to develop and validate AFM with specific antibodies (anti-TF and (b to compare air and liquid modes for EVs’ size and number determination as potential biomarkers of the prothrombotic risk. Methods: AFM multimode nanoscope III was used for air tapping mode (TM. AFM catalyst was used for liquid Peak Force Tapping (PFT mode. Vesicles are generated according to Davila et al.'s protocol. Substrates are coated with various concentrations of antibodies, thanks to ethanolamine and glutaraldehyde. Results: Vesicles were immobilised on antibody-coated surfaces to select tissue factor (TF-positive vesicles. The size range of vesicles observed in liquid PFT mode is 6–10 times higher than in air mode. This corresponds to the data found in the literature. Conclusion: We recommend liquid PFT mode to analyse vesicles on 5 µg/ml antibody-coated substrates.

  10. Temperature-dependent imaging of living cells by AFM

    International Nuclear Information System (INIS)

    Espenel, Cedric; Giocondi, Marie-Cecile; Seantier, Bastien; Dosset, Patrice; Milhiet, Pierre-Emmanuel; Le Grimellec, Christian

    2008-01-01

    Characterization of lateral organization of plasma membranes is a prerequisite to the understanding of membrane structure-function relationships in living cells. Lipid-lipid and lipid-protein interactions are responsible for the existence of various membrane microdomains involved in cell signalization and in numerous pathologies. Developing approaches for characterizing microdomains associate identification tools like recognition imaging with high-resolution topographical imaging. Membrane properties are markedly dependent on temperature. However, mesoscopic scale topographical information of cell surface in a temperature range covering most of cell biology experimentation is still lacking. In this work we have examined the possibility of imaging the temperature-dependent behavior of eukaryotic cells by atomic force microscopy (AFM). Our results establish that the surface of living CV1 kidney cells can be imaged by AFM, between 5 and 37 deg. C, both in contact and tapping modes. These first temperature-dependent data show that large cell structures appeared essentially stable at a microscopic scale. On the other hand, as shown by contact mode AFM, the surface was highly dynamic at a mesoscopic scale, with marked changes in apparent topography, friction, and deflection signals. When keeping the scanning conditions constant, a progressive loss in the image contrast was however observed, using tapping mode, on decreasing the temperature

  11. AFM measurements on ferrous materials

    Directory of Open Access Journals (Sweden)

    Ros Yáñez, T.

    2000-08-01

    Full Text Available The Atomic Force Microscope (AFM has been successfully used for the characterisation of the surface topography of a very wide range of materials. The AFM is an interesting technique to study the surface of materials and specifically to study surface effects in steel, e.g. after polishing and/or after etching. This technique allows not only to study the topography of the samples, but it is also a useful tool to obtain structural details of surfaces and for the identification of different phases. In this paper, some applications are shown concerning the study of the microstructure of multiphase-steels.

    El Microcopio de Fuerza Atómica (AFM es utilizado en la caracterización topográfica de una gran variedad de materiales. El AFM es una interesante técnica para el estudio de la superficie de los aceros, ya sea con ataque químico o no. Esta técnica es también una útil herramienta para obtener detalles estructurales en las superficies y en la identificación de fases. En este trabajo algunas de estas aplicaciones son mostradas en el estudio de la microestructura de aceros multifásicos.

  12. PREFACE: Non-contact AFM Non-contact AFM

    Science.gov (United States)

    Giessibl, Franz J.; Morita, Seizo

    2012-02-01

    This special issue is focussed on high resolution non-contact atomic force microscopy (AFM). Non-contact atomic force microscopy was established approximately 15 years ago as a tool to image conducting and insulating surfaces with atomic resolution. Since 1998, an annual international conference has taken place, and although the proceedings of these conferences are a useful source of information, several key developments warrant devoting a special issue to this subject. In the theoretic field, the possibility of supplementing established techniques such as scanning tunneling microscopy (STM) and Kelvin probe microscopy with atomically resolved force micrsoscopy poses many challenges in the calculation of contrast and contrast reversal. The surface science of insulators, self-assembled monolayers and adsorbates on insulators is a fruitful field for the application of non-contact AFM: several articles in this issue are devoted to these subjects. Atomic imaging and manipulation have been pioneered using STM, but because AFM allows the measurement of forces, AFM has had a profound impact in this field as well. Three-dimensional force spectroscopy has allowed many important insights into surface science. In this issue a combined 3D tunneling and force microscopy is introduced. Non-contact AFM typically uses frequency modulation to measure force gradients and was initially used mainly in a vacuum. As can be seen in this issue, frequency modulation is now also used in ambient conditions, allowing better spatial and force resolution. We thank all of the contributors for their time and efforts in making this special issue possible. We are also very grateful to the staff of IOP Publishing for handling the administrative aspects and for steering the refereeing process. Non-contact AFM contents Relation between the chemical force and the tunnelling current in atomic point contacts: a simple model Pavel Jelínek, Martin Ondrácek and Fernando Flores Theoretical simulation of

  13. Nano Mechanical Machining Using AFM Probe

    Science.gov (United States)

    Mostofa, Md. Golam

    Complex miniaturized components with high form accuracy will play key roles in the future development of many products, as they provide portability, disposability, lower material consumption in production, low power consumption during operation, lower sample requirements for testing, and higher heat transfer due to their very high surface-to-volume ratio. Given the high market demand for such micro and nano featured components, different manufacturing methods have been developed for their fabrication. Some of the common technologies in micro/nano fabrication are photolithography, electron beam lithography, X-ray lithography and other semiconductor processing techniques. Although these methods are capable of fabricating micro/nano structures with a resolution of less than a few nanometers, some of the shortcomings associated with these methods, such as high production costs for customized products, limited material choices, necessitate the development of other fabricating techniques. Micro/nano mechanical machining, such an atomic force microscope (AFM) probe based nano fabrication, has, therefore, been used to overcome some the major restrictions of the traditional processes. This technique removes material from the workpiece by engaging micro/nano size cutting tool (i.e. AFM probe) and is applicable on a wider range of materials compared to the photolithographic process. In spite of the unique benefits of nano mechanical machining, there are also some challenges with this technique, since the scale is reduced, such as size effects, burr formations, chip adhesions, fragility of tools and tool wear. Moreover, AFM based machining does not have any rotational movement, which makes fabrication of 3D features more difficult. Thus, vibration-assisted machining is introduced into AFM probe based nano mechanical machining to overcome the limitations associated with the conventional AFM probe based scratching method. Vibration-assisted machining reduced the cutting forces

  14. An AFM-SIMS Nano Tomography Acquisition System

    Science.gov (United States)

    Swinford, Richard William

    An instrument, adding the capability to measure 3D volumetric chemical composition, has been constructed by me as a member of the Sanchez Nano Laboratory. The laboratory's in situ atomic force microscope (AFM) and secondary ion mass spectrometry systems (SIMS) are functional and integrated as one instrument. The SIMS utilizes a Ga focused ion beam (FIB) combined with a quadrupole mass analyzer. The AFM is comprised of a 6-axis stage, three coarse axes and three fine. The coarse stage is used for placing the AFM tip anywhere inside a (13x13x5 mm3) (xyz) volume. Thus the tip can be moved in and out of the FIB processing region with ease. The planned range for the Z-axis piezo was 60 microm, but was reduced after it was damaged from arc events. The repaired Z-axis piezo is now operated at a smaller nominal range of 18 microm (16.7 microm after pre-loading), still quite respectable for an AFM. The noise floor of the AFM is approximately 0.4 nm Rq. The voxel size for the combined instrument is targeted at 50 nm or larger. Thus 0.4 nm of xyz uncertainty is acceptable. The instrument has been used for analyzing samples using FIB beam currents of 250 pA and 5.75 nA. Coarse tip approaches can take a long time so an abbreviated technique is employed. Because of the relatively long thro of the Z piezo, the tip can be disengaged by deactivating the servo PID. Once disengaged, it can be moved laterally out of the way of the FIB-SIMS using the coarse stage. This instrument has been used to acquire volumetric data on AlTiC using AFM tip diameters of 18.9 nm and 30.6 nm. Acquisition times are very long, requiring multiple days to acquire a 50-image stack. New features to be added include auto stigmation, auto beam shift, more software automation, etc. Longer term upgrades to include a new lower voltage Z-piezo with strain-gauge feedback and a new design to extend the life for the coarse XY nano-positioners. This AFM-SIMS instrument, as constructed, has proven to be a great proof

  15. Ultrasonically synthesized organic liquid-filled chitosan microcapsules: part 2: characterization using AFM (atomic force microscopy) and combined AFM-confocal laser scanning fluorescence microscopy.

    Science.gov (United States)

    Mettu, Srinivas; Ye, Qianyu; Zhou, Meifang; Dagastine, Raymond; Ashokkumar, Muthupandian

    2018-04-25

    Atomic Force Microscopy (AFM) is used to measure the stiffness and Young's modulus of individual microcapsules that have a chitosan cross-linked shell encapsulating tetradecane. The oil filled microcapsules were prepared using a one pot synthesis via ultrasonic emulsification of tetradecane and crosslinking of the chitosan shell in aqueous solutions of acetic acid. The concentration of acetic acid in aqueous solutions of chitosan was varied from 0.2% to 25% v/v. The effect of acetic acid concentration and size of the individual microcapsules on the strength was probed. The deformations and forces required to rupture the microcapsules were also measured. Three dimensional deformations of microcapsules under large applied loads were obtained by the combination of Laser Scanning Confocal Microscopy (LSCM) with Atomic Force Microscopy (AFM). The stiffness, and hence the modulus, of the microcapsules was found to decrease with an increase in size with the average stiffness ranging from 82 to 111 mN m-1 and average Young's modulus ranging from 0.4 to 6.5 MPa. The forces required to rupture the microcapsules varied from 150 to 250 nN with deformations of the microcapsules up to 62 to 110% relative to their radius, respectively. Three dimensional images obtained using laser scanning confocal microscopy showed that the microcapsules retained their structure and shape after being subjected to large deformations and subsequent removal of the loads. Based on the above observations, the oil filled chitosan crosslinked microcapsules are an ideal choice for use in the food and pharmaceutical industries as they would be able to withstand the process conditions encountered.

  16. High accuracy FIONA-AFM hybrid imaging

    International Nuclear Information System (INIS)

    Fronczek, D.N.; Quammen, C.; Wang, H.; Kisker, C.; Superfine, R.; Taylor, R.; Erie, D.A.; Tessmer, I.

    2011-01-01

    Multi-protein complexes are ubiquitous and play essential roles in many biological mechanisms. Single molecule imaging techniques such as electron microscopy (EM) and atomic force microscopy (AFM) are powerful methods for characterizing the structural properties of multi-protein and multi-protein-DNA complexes. However, a significant limitation to these techniques is the ability to distinguish different proteins from one another. Here, we combine high resolution fluorescence microscopy and AFM (FIONA-AFM) to allow the identification of different proteins in such complexes. Using quantum dots as fiducial markers in addition to fluorescently labeled proteins, we are able to align fluorescence and AFM information to ≥8 nm accuracy. This accuracy is sufficient to identify individual fluorescently labeled proteins in most multi-protein complexes. We investigate the limitations of localization precision and accuracy in fluorescence and AFM images separately and their effects on the overall registration accuracy of FIONA-AFM hybrid images. This combination of the two orthogonal techniques (FIONA and AFM) opens a wide spectrum of possible applications to the study of protein interactions, because AFM can yield high resolution (5-10 nm) information about the conformational properties of multi-protein complexes and the fluorescence can indicate spatial relationships of the proteins in the complexes. -- Research highlights: → Integration of fluorescent signals in AFM topography with high (<10 nm) accuracy. → Investigation of limitations and quantitative analysis of fluorescence-AFM image registration using quantum dots. → Fluorescence center tracking and display as localization probability distributions in AFM topography (FIONA-AFM). → Application of FIONA-AFM to a biological sample containing damaged DNA and the DNA repair proteins UvrA and UvrB conjugated to quantum dots.

  17. Simultaneous AFM and fluorescence imaging: A method for aligning an AFM-tip with an excitation beam using a 2D galvanometer

    Science.gov (United States)

    Moores, A. N.; Cadby, A. J.

    2018-02-01

    Correlative fluorescence and atomic force microscopy (AFM) imaging is a highly attractive technique for use in biological imaging, enabling force and mechanical measurements of particular structures whose locations are known due to the specificity of fluorescence imaging. The ability to perform these two measurements simultaneously (rather than consecutively with post-processing correlation) is highly valuable because it would allow the mechanical properties of a structure to be tracked over time as changes in the sample occur. We present an instrument which allows simultaneous AFM and fluorescence imaging by aligning an incident fluorescence excitation beam with an AFM-tip. Alignment was performed by calibrating a 2D galvanometer present in the excitation beam path and using it to reposition the incident beam. Two programs were developed (one manual and one automated) which correlate sample features between the AFM and fluorescence images, calculating the distance required to translate the incident beam towards the AFM-tip. Using this method, we were able to obtain beam-tip alignment (and therefore field-of-view alignment) from an offset of >15 μm to within one micron in two iterations of the program. With the program running alongside data acquisition for real-time feedback between AFM and optical images, this offset was maintained over a time period of several hours. Not only does this eliminate the need to image large areas with both techniques to ensure that fields-of-view overlap, but it also raises the possibility of using this instrument for tip-enhanced fluorescence applications, a technique in which super-resolution images have previously been achieved.

  18. LET spectrum measurements in Cr-39 PNTD with AFM

    Energy Technology Data Exchange (ETDEWEB)

    Johnson, Carl Edward [Los Alamos National Laboratory; De Witt, Joel M [OSU, PHYSICS; Benton, Eric R [OSU, PHYSICS; Yasuda, Nakahiro [NIRS, HIMAC; Benton, Eugene V [UNIV OF SAN FRANCISCO

    2010-01-01

    Energetic protons, neutrons, and heavy ions undergoing collisions with target nuclei of varying Z can produce residual heavy recoil fragments via intra-nuclear cascade/evaporation reactions. The particles produced in these non-elastic collisions generally have such extremely short range ({approx}< 10 {mu}m) that they cannot be directly observed by conventional detection methods including CR-39 plastic nuclear track detector (PNTD) that has been chemically etched for analysis by standard visible light microscopy. However, high-LET recoil fragments having range on the order of several cell diameters can be produced in tissue during radiotherapy using proton and carbon beams. We have developed a method to analyze short-range, high-LET tracks in CR-39 plastic nuclear track detector (PNTD) using short duration chemical etching ({approx}< 1 {mu}m) followed by automated atomic force microscope (AFM) scanning. The post-scan data processing used in this work was based on semi-automated matrix analysis opposed to traditional grey-scale image analysis. This method takes advantage of the 3-D data obtained via AFM to achieve robust discrimination of nuclear tracks from other features. Through automation of AFM scanning, sufficient AFM scan frames were obtained to attain an LET spectrum spanning the LET range from 200-1500 keV/{mu}m. In addition to our experiments, simulations were carried out with the Monte Carlo transport code, FLUKA. To demonstrate this method, CR-39 PNTD was exposed to the proton therapy beam at Loma Linda University Medical Center (LLUMC) at 60 and 230 MeV. Additionally, detectors were exposed to I GeV protons at the NASA Space Radiation Laboratory (NSRL) at Brookhaven National Laboratory (BNL). For these exposures CR-39 PNTD, Al and Cu target foils were used between detector layers.

  19. LET spectrum measurements in Cr-39 PNTD with AFM

    International Nuclear Information System (INIS)

    Johnson, Carl Edward; DeWitt, Joel M.; Benton, Eric R.; Yasuda, Nakahiro; Benton, Eugene V.

    2010-01-01

    Energetic protons, neutrons, and heavy ions undergoing collisions with target nuclei of varying Z can produce residual heavy recoil fragments via intra-nuclear cascade/evaporation reactions. The particles produced in these non-elastic collisions generally have such extremely short range (∼< 10 μm) that they cannot be directly observed by conventional detection methods including CR-39 plastic nuclear track detector (PNTD) that has been chemically etched for analysis by standard visible light microscopy. However, high-LET recoil fragments having range on the order of several cell diameters can be produced in tissue during radiotherapy using proton and carbon beams. We have developed a method to analyze short-range, high-LET tracks in CR-39 plastic nuclear track detector (PNTD) using short duration chemical etching (∼< 1 μm) followed by automated atomic force microscope (AFM) scanning. The post-scan data processing used in this work was based on semi-automated matrix analysis opposed to traditional grey-scale image analysis. This method takes advantage of the 3-D data obtained via AFM to achieve robust discrimination of nuclear tracks from other features. Through automation of AFM scanning, sufficient AFM scan frames were obtained to attain an LET spectrum spanning the LET range from 200-1500 keV/μm. In addition to our experiments, simulations were carried out with the Monte Carlo transport code, FLUKA. To demonstrate this method, CR-39 PNTD was exposed to the proton therapy beam at Loma Linda University Medical Center (LLUMC) at 60 and 230 MeV. Additionally, detectors were exposed to I GeV protons at the NASA Space Radiation Laboratory (NSRL) at Brookhaven National Laboratory (BNL). For these exposures CR-39 PNTD, Al and Cu target foils were used between detector layers.

  20. Cell visco-elasticity measured with AFM and optical trapping at sub-micrometer deformations.

    Directory of Open Access Journals (Sweden)

    Schanila Nawaz

    Full Text Available The measurement of the elastic properties of cells is widely used as an indicator for cellular changes during differentiation, upon drug treatment, or resulting from the interaction with the supporting matrix. Elasticity is routinely quantified by indenting the cell with a probe of an AFM while applying nano-Newton forces. Because the resulting deformations are in the micrometer range, the measurements will be affected by the finite thickness of the cell, viscous effects and even cell damage induced by the experiment itself. Here, we have analyzed the response of single 3T3 fibroblasts that were indented with a micrometer-sized bead attached to an AFM cantilever at forces from 30-600 pN, resulting in indentations ranging from 0.2 to 1.2 micrometer. To investigate the cellular response at lower forces up to 10 pN, we developed an optical trap to indent the cell in vertical direction, normal to the plane of the coverslip. Deformations of up to two hundred nanometers achieved at forces of up to 30 pN showed a reversible, thus truly elastic response that was independent on the rate of deformation. We found that at such small deformations, the elastic modulus of 100 Pa is largely determined by the presence of the actin cortex. At higher indentations, viscous effects led to an increase of the apparent elastic modulus. This viscous contribution that followed a weak power law, increased at larger cell indentations. Both AFM and optical trapping indentation experiments give consistent results for the cell elasticity. Optical trapping has the benefit of a lower force noise, which allows a more accurate determination of the absolute indentation. The combination of both techniques allows the investigation of single cells at small and large indentations and enables the separation of their viscous and elastic components.

  1. A scanning probe microscope for magnetoresistive cantilevers utilizing a nested scanner design for large-area scans

    Directory of Open Access Journals (Sweden)

    Tobias Meier

    2015-02-01

    Full Text Available We describe an atomic force microscope (AFM for the characterization of self-sensing tunneling magnetoresistive (TMR cantilevers. Furthermore, we achieve a large scan-range with a nested scanner design of two independent piezo scanners: a small high resolution scanner with a scan range of 5 × 5 × 5 μm3 is mounted on a large-area scanner with a scan range of 800 × 800 × 35 μm3. In order to characterize TMR sensors on AFM cantilevers as deflection sensors, the AFM is equipped with a laser beam deflection setup to measure the deflection of the cantilevers independently. The instrument is based on a commercial AFM controller and capable to perform large-area scanning directly without stitching of images. Images obtained on different samples such as calibration standard, optical grating, EPROM chip, self-assembled monolayers and atomic step-edges of gold demonstrate the high stability of the nested scanner design and the performance of self-sensing TMR cantilevers.

  2. A concept for automated nanoscale atomic force microscope (AFM) measurements using a priori knowledge

    International Nuclear Information System (INIS)

    Recknagel, C; Rothe, H

    2009-01-01

    The nanometer coordinate measuring machine (NCMM) is developed for comparatively fast large area scans with high resolution. The system combines a metrological atomic force microscope (AFM) with a precise positioning system. The sample is moved under the probe system via the positioning system achieving a scan range of 25 × 25 × 5 mm 3 with a resolution of 0.1 nm. A concept for AFM measurements using a priori knowledge is implemented. The a priori knowledge is generated through measurements with a white light interferometer and the use of CAD data. Dimensional markup language is used as a transfer and target format for a priori knowledge and measurement data. Using the a priori knowledge and template matching algorithms combined with the optical microscope of the NCMM, the region of interest can automatically be identified. In the next step the automatic measurement of the part coordinate system and the measurement elements with the AFM sensor of the NCMM is done. The automatic measurement involves intelligent measurement strategies, which are adapted to specific geometries of the measurement feature to reduce measurement time and drift effects

  3. Optimization of specimen preparation of thin cell section for AFM observation

    Energy Technology Data Exchange (ETDEWEB)

    Li Xinhui [Nanobiology Laboratory, Bio-X Life Science Research Center, School of Life Science and Biotechnology, Shanghai Jiao Tong University, 800 Dongchuan Road, Minhang District, Shanghai 200240 (China); Ji Tong [Department of Oral and Maxillofacial Surgery, School of Stomatology, Affiliated Ninth People' s Hospital, Medical School, Shanghai Jiao Tong University, Shanghai 200011 (China); Hu Jun [Nanobiology Laboratory, Bio-X Life Science Research Center, School of Life Science and Biotechnology, Shanghai Jiao Tong University, 800 Dongchuan Road, Minhang District, Shanghai 200240 (China); Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai 201800 (China); Sun Jielin [Nanobiology Laboratory, Bio-X Life Science Research Center, School of Life Science and Biotechnology, Shanghai Jiao Tong University, 800 Dongchuan Road, Minhang District, Shanghai 200240 (China)], E-mail: jlsun@sjtu.edu.cn

    2008-08-15

    High resolution imaging of intracellular structures of ultrathin cell section samples is critical to the performance of precise manipulation by atomic force microscopy (AFM). Here, we test the effect of multiple factors during section sample preparation on the quality of the AFM image. These factors include the embedding materials, the annealing process of the specimen block, section thickness, and section side. We found that neither the embedding materials nor the temperature and speed of the annealing process has any effect on AFM image resolution. However, the section thickness and section side significantly affect the surface topography and AFM image resolution. By systematically testing the image quality of both sides of cell sections over a wide range of thickness (40-1000 nm), we found that the best resolution was obtained with upper-side sections approximately 50-100 nm thick. With these samples, we could observe precise structure details of the cell, including its membrane, nucleoli, and other organelles. Similar results were obtained for other cell types, including Tca8113, C6, and ECV-304. In brief, by optimizing the condition of ultrathin cell section preparation, we were able to obtain high resolution intracellular AFM images, which provide an essential basis for further AFM manipulation.

  4. Laser-damage susceptibility of nodular defects in dielectric mirror coatings: AFM measurements and electric-field modeling

    International Nuclear Information System (INIS)

    Kozlowski, M.R.; DeFord, J.F.; Staggs, M.C.

    1993-01-01

    Atomic force microscopy (AFM) and electromagnetic field modeling were used to study the influence of nodular coating defects on laser-induced damage of multilayer dielectric coatings. In studies of HfO 2 /SiO 2 mirrors with 1.06 μm illumination, AFM results showed that nodular defects with high dome heights (>0.6 μm) were most susceptible to laser damage. Crater defects, formed by nodules ejected from the coating prior to illumination, were not damaged when illuminated over the same range of fluences. A finite-difference time-domain electromagnetic modeling code was used to study the influence of 3-D nodule defects on the E-field distribution within the interference coating. The modeling results show that Enfield enhancements as large as a factor of 4 can be present at the defects. Crater defects, however, result in minimal enhancement of the E-fields within the coating. These modeling results are consistent with the AFM experimental data, indicating that E-field enhancement is a contributing mechanism in defect-dominated laser damage of optical coatings

  5. Measuring bacterial cells size with AFM

    Directory of Open Access Journals (Sweden)

    Denise Osiro

    2012-03-01

    Full Text Available Atomic Force Microscopy (AFM can be used to obtain high-resolution topographical images of bacteria revealing surface details and cell integrity. During scanning however, the interactions between the AFM probe and the membrane results in distortion of the images. Such distortions or artifacts are the result of geometrical effects related to bacterial cell height, specimen curvature and the AFM probe geometry. The most common artifact in imaging is surface broadening, what can lead to errors in bacterial sizing. Several methods of correction have been proposed to compensate for these artifacts and in this study we describe a simple geometric model for the interaction between the tip (a pyramidal shaped AFM probe and the bacterium (Escherichia coli JM-109 strain to minimize the enlarging effect. Approaches to bacteria immobilization and examples of AFM images analysis are also described.

  6. Nonlinear Phenomena in the Single-Mode Dynamics in an AFM Cantilever Beam

    KAUST Repository

    Ruzziconi, Laura

    2016-12-05

    This study deals with the nonlinear dynamics arising in an atomic force microscope cantilever beam. After analyzing the static behavior, a single degree of freedom Galerkin reduced order model is introduced, which describes the overall scenario of the structure response in a neighborhood of the primary resonance. Extensive numerical simulations are performed when both the forcing amplitude and frequency are varied, ranging from low up to elevated excitations. The coexistence of competing attractors with different characteristics is analyzed. Both the non-resonant and the resonant behavior are observed, as well as ranges of inevitable escape. Versatility of behavior is highlighted, which may be attractive in applications. Special attention is devoted to the effects of the tip-sample separation distance, since this aspect is of fundamental importance to understand the operation of an AFM. We explore the metamorphoses of the multistability region when the tip-sample separation distance is varied. To have a complete description of the AFM response, comprehensive behavior charts are introduced to detect the theoretical boundaries of appearance and disappearance of the main attractors. Also, extensive numerical simulations investigate the AFM response when both the forcing amplitude and the tip-sample separation distance are considered as control parameters. The main features are analyzed in detail and the obtained results are interpreted in terms of oscillations of the cantilever-tip ensemble. However, we note that all the aforementioned results represent the limit when disturbances are absent, which never occurs in practice. Here comes the importance of overcoming local investigations and exploring dynamics from a global perspective, by introducing dynamical integrity concepts. To extend the AFM results to the practical case where disturbances exist, we develop a dynamical integrity analysis. After performing a systematic basin of attraction analysis, integrity

  7. High-speed AFM for Studying Dynamic Biomolecular Processes

    Science.gov (United States)

    Ando, Toshio

    2008-03-01

    Biological molecules show their vital activities only in aqueous solutions. It had been one of dreams in biological sciences to directly observe biological macromolecules (protein, DNA) at work under a physiological condition because such observation is straightforward to understanding their dynamic behaviors and functional mechanisms. Optical microscopy has no sufficient spatial resolution and electron microscopy is not applicable to in-liquid samples. Atomic force microscopy (AFM) can visualize molecules in liquids at high resolution but its imaging rate was too low to capture dynamic biological processes. This slow imaging rate is because AFM employs mechanical probes (cantilevers) and mechanical scanners to detect the sample height at each pixel. It is quite difficult to quickly move a mechanical device of macroscopic size with sub-nanometer accuracy without producing unwanted vibrations. It is also difficult to maintain the delicate contact between a probe tip and fragile samples. Two key techniques are required to realize high-speed AFM for biological research; fast feedback control to maintain a weak tip-sample interaction force and a technique to suppress mechanical vibrations of the scanner. Various efforts have been carried out in the past decade to materialize high-speed AFM. The current high-speed AFM can capture images on video at 30-60 frames/s for a scan range of 250nm and 100 scan lines, without significantly disturbing week biomolecular interaction. Our recent studies demonstrated that this new microscope can reveal biomolecular processes such as myosin V walking along actin tracks and association/dissociation dynamics of chaperonin GroEL-GroES that occurs in a negatively cooperative manner. The capacity of nanometer-scale visualization of dynamic processes in liquids will innovate on biological research. In addition, it will open a new way to study dynamic chemical/physical processes of various phenomena that occur at the liquid-solid interfaces.

  8. The AFm phase in Portland cement

    International Nuclear Information System (INIS)

    Matschei, T.; Lothenbach, B.; Glasser, F.P.

    2007-01-01

    The AFm phase of Portland cements refers to a family of hydrated calcium aluminates based on the hydrocalumite-like structure of 4CaO.Al 2 O 3 .13-19 H 2 O. However OH - may be replaced by SO 4 2- and CO 3 2- . Except for limited replacement (50 mol%, maximum) of sulfate by hydroxide, these compositions do not form solid solutions and, from the mineralogical standpoint, behave as separate phases. Therefore many hydrated cements will contain mixtures of AFm phases. AFm phases have been made from precursors and experimentally-determined phase relationships are depicted at 25 deg. C. Solubility data are reported and thermodynamic data are derived. The 25 deg. C stability of AFm phases is much affected by the nature of the anion: carbonate stabilises AFm and displaces OH and SO 4 at species activities commonly encountered in cement systems. However in the presence of portlandite, and as carbonate displaces sulfate in AFm, the reaction results in changes in the amount of both portlandite and ettringite: specimen calculations are presented to quantify these changes. The scheme of phase balances enables calculation of the mineralogical balances of a hydrated cement paste with greater accuracy than hitherto practicable

  9. AFM annual report 1983 (Petroleum Industry). [Export Federation for Mineral Oil, Federal Republic of Germany]. AFM Jahresbericht 1983

    Energy Technology Data Exchange (ETDEWEB)

    1983-01-01

    The annual report 1983 of the Export Federation for Mineral Oil (AFM) contains informations about the mineral oil economics, the market development for selected main products and the environmental protection. The AFM terms (standard conditions for barge transactions) for the mineral oil industry are given. The AFM Oil Market Report Daily has extended the frame of its reports in 1983.

  10. Reconstruction Algorithms in Undersampled AFM Imaging

    DEFF Research Database (Denmark)

    Arildsen, Thomas; Oxvig, Christian Schou; Pedersen, Patrick Steffen

    2016-01-01

    This paper provides a study of spatial undersampling in atomic force microscopy (AFM) imaging followed by different image reconstruction techniques based on sparse approximation as well as interpolation. The main reasons for using undersampling is that it reduces the path length and thereby...... the scanning time as well as the amount of interaction between the AFM probe and the specimen. It can easily be applied on conventional AFM hardware. Due to undersampling, it is then necessary to further process the acquired image in order to reconstruct an approximation of the image. Based on real AFM cell...... images, our simulations reveal that using a simple raster scanning pattern in combination with conventional image interpolation performs very well. Moreover, this combination enables a reduction by a factor 10 of the scanning time while retaining an average reconstruction quality around 36 dB PSNR...

  11. A comparison of atomic force microscopy (AFM) and dynamic light scattering (DLS) methods to characterize nanoparticle size distributions

    International Nuclear Information System (INIS)

    Hoo, Christopher M.; Starostin, Natasha; West, Paul; Mecartney, Martha L.

    2008-01-01

    This paper compares the accuracy of conventional dynamic light scattering (DLS) and atomic force microscopy (AFM) for characterizing size distributions of polystyrene nanoparticles in the size range of 20-100 nm. Average DLS values for monosize dispersed particles are slightly higher than the nominal values whereas AFM values were slightly lower than nominal values. Bimodal distributions were easily identified with AFM, but DLS results were skewed toward larger particles. AFM characterization of nanoparticles using automated analysis software provides an accurate and rapid analysis for nanoparticle characterization and has advantages over DLS for non-monodispersed solutions.

  12. Surface analysis with STM and AFM

    CERN Document Server

    Magonov, Sergi N

    1996-01-01

    Scanning tunneling microscopy (STM) and atomic force microscopy (AFM) are powerful tools for surface examination. In the past, many STM and AFM studies led to erroneous conclusions due to lack of proper theoretical considerations and of an understanding of how image patterns are affected by measurement conditions. For this book, two world experts, one on theoretical analysis and the other on experimental characterization, have joined forces to bring together essential components of STM and AFM studies: The practical aspects of STM, the image simulation by surface electron density plot calculat

  13. AFM tip characterization by using FFT filtered images of step structures

    Energy Technology Data Exchange (ETDEWEB)

    Yan, Yongda, E-mail: yanyongda@hit.edu.cn [Key Laboratory of Micro-systems and Micro-structures Manufacturing of Ministry of Education, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Center For Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Xue, Bo [Key Laboratory of Micro-systems and Micro-structures Manufacturing of Ministry of Education, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Center For Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Hu, Zhenjiang; Zhao, Xuesen [Center For Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China)

    2016-01-15

    The measurement resolution of an atomic force microscope (AFM) is largely dependent on the radius of the tip. Meanwhile, when using AFM to study nanoscale surface properties, the value of the tip radius is needed in calculations. As such, estimation of the tip radius is important for analyzing results taken using an AFM. In this study, a geometrical model created by scanning a step structure with an AFM tip was developed. The tip was assumed to have a hemispherical cone shape. Profiles simulated by tips with different scanning radii were calculated by fast Fourier transform (FFT). By analyzing the influence of tip radius variation on the spectra of simulated profiles, it was found that low-frequency harmonics were more susceptible, and that the relationship between the tip radius and the low-frequency harmonic amplitude of the step structure varied monotonically. Based on this regularity, we developed a new method to characterize the radius of the hemispherical tip. The tip radii estimated with this approach were comparable to the results obtained using scanning electron microscope imaging and blind reconstruction methods. - Highlights: • The AFM tips with different radii were simulated to scan a nano-step structure. • The spectra of the simulation scans under different radii were analyzed. • The functions of tip radius and harmonic amplitude were used for evaluating tip. • The proposed method has been validated by SEM imaging and blind reconstruction.

  14. Measuring protein isoelectric points by AFM-based force spectroscopy using trace amounts of sample

    Science.gov (United States)

    Guo, Shifeng; Zhu, Xiaoying; Jańczewski, Dominik; Lee, Serina Siew Chen; He, Tao; Teo, Serena Lay Ming; Vancso, G. Julius

    2016-09-01

    Protein charge at various pH and isoelectric point (pI) values is important in understanding protein function. However, often only trace amounts of unknown proteins are available and pI measurements cannot be obtained using conventional methods. Here, we show a method based on the atomic force microscope (AFM) to determine pI using minute quantities of proteins. The protein of interest is immobilized on AFM colloidal probes and the adhesion force of the protein is measured against a positively and a negatively charged substrate made by layer-by-layer deposition of polyelectrolytes. From the AFM force-distance curves, pI values with an estimated accuracy of ±0.25 were obtained for bovine serum albumin, myoglobin, fibrinogen and ribonuclease A over a range of 4.7-9.8. Using this method, we show that the pI of the ‘footprint’ of the temporary adhesive proteins secreted by the barnacle cyprid larvae of Amphibalanus amphitrite is in the range 9.6-9.7.

  15. Fabrication and buckling dynamics of nanoneedle AFM probes

    Energy Technology Data Exchange (ETDEWEB)

    Beard, J D; Gordeev, S N, E-mail: jdb28@bath.ac.uk [Department of Physics, University of Bath, Claverton Down, Bath BA2 7AY (United Kingdom)

    2011-04-29

    A new method for the fabrication of high-aspect-ratio probes by electron beam induced deposition is described. This technique allows the fabrication of cylindrical 'nanoneedle' structures on the atomic force microscope (AFM) probe tip which can be used for accurate imaging of surfaces with high steep features. Scanning electron microscope (SEM) imaging showed that needles with diameters in the range of 18-100 nm could be obtained by this technique. The needles were shown to undergo buckling deformation under large tip-sample forces. The deformation was observed to recover elastically under vertical deformations of up to {approx} 60% of the needle length, preventing damage to the needle. A technique of stabilizing the needle against buckling by coating it with additional electron beam deposited carbon was also investigated; it was shown that coated needles of 75 nm or greater total diameter did not buckle even under tip-sample forces of {approx} 1.5 {mu}N.

  16. BOREAS AFM-11 Aircraft Flux Analysis Reports

    Data.gov (United States)

    National Aeronautics and Space Administration — Reports from the BOREAS AFM-11 team regarding quality control and sampling analysis of data collected by other AFM personnel using the Electra, LongEZ, and Twin...

  17. Quantitative Assessment of Aflatoxin (AFM1) in Milk Collected from Dairy Farms in Faisalabad, Pakistan

    International Nuclear Information System (INIS)

    Sajid, M.W.; Randhawa, M.A.; Zahoor, T.; Sultan, J.I.

    2015-01-01

    Milk contamination with aflatoxin (AFM1) is an issue of great concern in developing countries like Pakistan which demands a great attention. Milk constitutes an important part of human diet, particularly for the youngs. So, it is our utmost need to assess the presence of AFM1 in milk. In the present study assessment of AFM1 in milk collected from different dairy farms of Faisalabad was carried out using High Performance Liquid Chromatography (HPLC) equipped with Fluorescence detector. The results were compared with pre-established maximum residual limit (MRL) in order to evaluate the safety of milk for human consumption. The study revealed that all the 50 tested samples were found positive for AFM1 contamination at various levels. Among buffalo dairy farms concentration of AFM1 ranged between 0.0513 λg L-1 and 0.1006 μg L-1. From the cow dairy farms, the AFM1 contamination level was found lowest with a mean of 0.0397 μg L-1 and the highest AFM1 contamination level was with a mean of 0.1143 μg L-1. Overall percentage of AFM1 contamination and concentration levels were found higher in the milk collected from buffalo dairy farms as compared to cow dairy farms. 21 out of 25 (84 percentage) buffalo and 18 out of 25 (72 percentage) cow milk samples were exceeded the European Commission MRL of 0.050 mu g L-1. The results of the present study will be helpful for regulations implementation in order to minimize or avoid the AFM1 contamination in milk from the farms in the study area. (author)

  18. Amyloid and membrane complexity: The toxic interplay revealed by AFM.

    Science.gov (United States)

    Canale, Claudio; Oropesa-Nuñez, Reinier; Diaspro, Alberto; Dante, Silvia

    2018-01-01

    Lipid membranes play a fundamental role in the pathological development of protein misfolding diseases. Several pieces of evidence suggest that the lipid membrane could act as a catalytic surface for protein aggregation. Furthermore, a leading theory indicates the interaction between the cell membrane and misfolded oligomer species as the responsible for cytotoxicity, hence, for neurodegeneration in disorders such as Alzheimer's and Parkinson's disease. The definition of the mechanisms that drive the interaction between pathological protein aggregates and plasma membrane is fundamental for the development of effective therapies for a large class of diseases. Atomic force microscopy (AFM) has been employed to study how amyloid aggregates affect the cell physiological properties. Considerable efforts were spent to characterize the interaction with model systems, i.e., planar supported lipid bilayers, but some works also addressed the problem directly on living cells. Here, an overview of the main works involving the use of the AFM on both model system and living cells will be provided. Different kind of approaches will be presented, as well as the main results derived from the AFM analysis. Copyright © 2017 Elsevier Ltd. All rights reserved.

  19. SU-8 hollow cantilevers for AFM cell adhesion studies

    Science.gov (United States)

    Martinez, Vincent; Behr, Pascal; Drechsler, Ute; Polesel-Maris, Jérôme; Potthoff, Eva; Vörös, Janos; Zambelli, Tomaso

    2016-05-01

    A novel fabrication method was established to produce flexible, transparent, and robust tipless hollow atomic force microscopy (AFM) cantilevers made entirely from SU-8. Channels of 3 μm thickness and several millimeters length were integrated into 12 μm thick and 40 μm wide cantilevers. Connected to a pressure controller, the devices showed high sealing performance with no leakage up to 6 bars. Changing the cantilever lengths from 100 μm to 500 μm among the same wafer allowed the targeting of various spring constants ranging from 0.5 to 80 N m-1 within a single fabrication run. These hollow polymeric AFM cantilevers were operated in the optical beam deflection configuration. To demonstrate the performance of the device, single-cell force spectroscopy experiments were performed with a single probe detaching in a serial protocol more than 100 Saccharomyces cerevisiae yeast cells from plain glass and glass coated with polydopamine while measuring adhesion forces in the sub-nanoNewton range. SU-8 now offers a new alternative to conventional silicon-based hollow cantilevers with more flexibility in terms of complex geometric design and surface chemistry modification.

  20. SU-8 hollow cantilevers for AFM cell adhesion studies

    International Nuclear Information System (INIS)

    Martinez, Vincent; Behr, Pascal; Vörös, Janos; Zambelli, Tomaso; Drechsler, Ute; Polesel-Maris, Jérôme; Potthoff, Eva

    2016-01-01

    A novel fabrication method was established to produce flexible, transparent, and robust tipless hollow atomic force microscopy (AFM) cantilevers made entirely from SU-8. Channels of 3 μm thickness and several millimeters length were integrated into 12 μm thick and 40 μm wide cantilevers. Connected to a pressure controller, the devices showed high sealing performance with no leakage up to 6 bars. Changing the cantilever lengths from 100 μm to 500 μm among the same wafer allowed the targeting of various spring constants ranging from 0.5 to 80 N m −1 within a single fabrication run. These hollow polymeric AFM cantilevers were operated in the optical beam deflection configuration. To demonstrate the performance of the device, single-cell force spectroscopy experiments were performed with a single probe detaching in a serial protocol more than 100 Saccharomyces cerevisiae yeast cells from plain glass and glass coated with polydopamine while measuring adhesion forces in the sub-nanoNewton range. SU-8 now offers a new alternative to conventional silicon-based hollow cantilevers with more flexibility in terms of complex geometric design and surface chemistry modification. (paper)

  1. Controlled AFM detachments and movement of nanoparticles: gold clusters on HOPG at different temperatures.

    Science.gov (United States)

    Tripathi, Manoj; Paolicelli, Guido; D'Addato, Sergio; Valeri, Sergio

    2012-06-22

    The effect of temperature on the onset of movement of gold nanoclusters (diameter 27 nm) deposited on highly oriented pyrolytic graphite (HOPG) has been studied by atomic force microscopy (AFM) techniques. Using the AFM with amplitude modulation (tapping mode AFM) we have stimulated and controlled the movement of individual clusters. We show how, at room temperature, controlled detachments and smooth movements can be obtained for clusters having dimensions comparable to or smaller than the tip radius. Displacement is practically visible in real time and it can be started and stopped easily by adjusting only one parameter, the tip amplitude oscillation. Analysing the energy dissipation signal at the onset of nanocluster sliding we evaluated a detachment threshold energy as a function of temperature in the range 300-413 K. We also analysed single cluster thermal induced displacement and combining this delicate procedure with AFM forced movement behaviour we conclude that detachment threshold energy is directly related to the activation energy of nanocluster diffusion and it scales linearly with temperature as expected for a single-particle thermally activated process.

  2. AFM image of an entire polygene chromosome

    International Nuclear Information System (INIS)

    Li Minqian; Takeuchi; Ikai, A.

    1994-01-01

    The author present AFM images of an entire polygene chromosome of Drosophila for the first time. Comparing with conventional optical microscope, the AFM image of the polygene chromosomes provides much higher resolution and 3-D measurement capability which will lead to finer scale gene mapping and identification

  3. MetaRep, an extended CMAS 3D program to visualize mafic (CMAS, ACF-S, ACF-N) and pelitic (AFM-K, AFM-S, AKF-S) projections

    Science.gov (United States)

    France, Lydéric; Nicollet, Christian

    2010-06-01

    MetaRep is a program based on our earlier program CMAS 3D. It is developed in MATLAB ® script. MetaRep objectives are to visualize and project major element compositions of mafic and pelitic rocks and their minerals in the pseudo-quaternary projections of the ACF-S, ACF-N, CMAS, AFM-K, AFM-S and AKF-S systems. These six systems are commonly used to describe metamorphic mineral assemblages and magmatic evolutions. Each system, made of four apices, can be represented in a tetrahedron that can be visualized in three dimensions with MetaRep; the four tetrahedron apices represent oxides or combination of oxides that define the composition of the projected rock or mineral. The three-dimensional representation allows one to obtain a better understanding of the topology of the relationships between the rocks and minerals and relations. From these systems, MetaRep can also project data in ternary plots (for example, the ACF, AFM and AKF ternary projections can be generated). A functional interface makes it easy to use and does not require any knowledge of MATLAB ® programming. To facilitate the use, MetaRep loads, from the main interface, data compiled in a Microsoft Excel ™ spreadsheet. Although useful for scientific research, the program is also a powerful tool for teaching. We propose an application example that, by using two combined systems (ACF-S and ACF-N), provides strong confirmation in the petrological interpretation.

  4. Accuracy optimization of high-speed AFM measurements using Design of Experiments

    DEFF Research Database (Denmark)

    Tosello, Guido; Marinello, F.; Hansen, Hans Nørgaard

    2010-01-01

    Atomic Force Microscopy (AFM) is being increasingly employed in industrial micro/nano manufacturing applications and integrated into production lines. In order to achieve reliable process and product control at high measuring speed, instrument optimization is needed. Quantitative AFM measurement...... results are influenced by a number of scan settings parameters, defining topography sampling and measurement time: resolution (number of profiles and points per profile), scan range and direction, scanning force and speed. Such parameters are influencing lateral and vertical accuracy and, eventually......, the estimated dimensions of measured features. The definition of scan settings is based on a comprehensive optimization that targets maximization of information from collected data and minimization of measurement uncertainty and scan time. The Design of Experiments (DOE) technique is proposed and applied...

  5. Quantitative analysis and classification of AFM images of human hair.

    Science.gov (United States)

    Gurden, S P; Monteiro, V F; Longo, E; Ferreira, M M C

    2004-07-01

    The surface topography of human hair, as defined by the outer layer of cellular sheets, termed cuticles, largely determines the cosmetic properties of the hair. The condition of the cuticles is of great cosmetic importance, but also has the potential to aid diagnosis in the medical and forensic sciences. Atomic force microscopy (AFM) has been demonstrated to offer unique advantages for analysis of the hair surface, mainly due to the high image resolution and the ease of sample preparation. This article presents an algorithm for the automatic analysis of AFM images of human hair. The cuticular structure is characterized using a series of descriptors, such as step height, tilt angle and cuticle density, allowing quantitative analysis and comparison of different images. The usefulness of this approach is demonstrated by a classification study. Thirty-eight AFM images were measured, consisting of hair samples from (a) untreated and bleached hair samples, and (b) the root and distal ends of the hair fibre. The multivariate classification technique partial least squares discriminant analysis is used to test the ability of the algorithm to characterize the images according to the properties of the hair samples. Most of the images (86%) were found to be classified correctly.

  6. Statistical analysis of AFM topographic images of self-assembled quantum dots

    Energy Technology Data Exchange (ETDEWEB)

    Sevriuk, V. A.; Brunkov, P. N., E-mail: brunkov@mail.ioffe.ru; Shalnev, I. V.; Gutkin, A. A.; Klimko, G. V.; Gronin, S. V.; Sorokin, S. V.; Konnikov, S. G. [Russian Academy of Sciences, Ioffe Physical-Technical Institute (Russian Federation)

    2013-07-15

    To obtain statistical data on quantum-dot sizes, AFM topographic images of the substrate on which the dots under study are grown are analyzed. Due to the nonideality of the substrate containing height differences on the order of the size of nanoparticles at distances of 1-10 {mu}m and the insufficient resolution of closely arranged dots due to the finite curvature radius of the AFM probe, automation of the statistical analysis of their large dot array requires special techniques for processing topographic images to eliminate the loss of a particle fraction during conventional processing. As such a technique, convolution of the initial matrix of the AFM image with a specially selected matrix is used. This makes it possible to determine the position of each nanoparticle and, using the initial matrix, to measure their geometrical parameters. The results of statistical analysis by this method of self-assembled InAs quantum dots formed on the surface of an AlGaAs epitaxial layer are presented. It is shown that their concentration, average size, and half-width of height distribution depend strongly on the In flow and total amount of deposited InAs which are varied within insignificant limits.

  7. Custom AFM for X-ray beamlines: in situ biological investigations under physiological conditions

    Energy Technology Data Exchange (ETDEWEB)

    Gumí-Audenis, B. [ESRF, The European Synchrotron, Grenoble (France); Institute for Bioengineering of Catalonia (IBEC), Barcelona (Spain); Physical Chemistry Department, Universitat de Barcelona, Barcelona (Spain); Networking Biomedical Research Center on Bioengineering, Biomaterials and Nanomedicine (CIBER-BBN), Madrid (Spain); Carlà, F. [ESRF, The European Synchrotron, Grenoble (France); Vitorino, M. V. [University of Lisboa, Falculty of Science, Biosystems and Integrative Sciences Institute - BIOISI, Lisbon (Portugal); Panzarella, A. [ESRF, The European Synchrotron, Grenoble (France); Porcar, L. [Institut Laue-Langevin, Grenoble (France); Boilot, M. [ORTEC, Marseille (France); Guerber, S. [CEA, LETI Grenoble (France); Bernard, P. [ESRF, The European Synchrotron, Grenoble (France); Rodrigues, M. S. [University of Lisboa, Falculty of Science, Biosystems and Integrative Sciences Institute - BIOISI, Lisbon (Portugal); Sanz, F.; Giannotti, M. I. [Institute for Bioengineering of Catalonia (IBEC), Barcelona (Spain); Physical Chemistry Department, Universitat de Barcelona, Barcelona (Spain); Networking Biomedical Research Center on Bioengineering, Biomaterials and Nanomedicine (CIBER-BBN), Madrid (Spain); Costa, L., E-mail: luca.costa@esrf.fr [ESRF, The European Synchrotron, Grenoble (France)

    2015-09-30

    The performance of a custom atomic force microscope for grazing-incidence X-ray experiments on hydrated soft and biological samples is presented. A fast atomic force microscope (AFM) has been developed that can be installed as a sample holder for grazing-incidence X-ray experiments at solid/gas or solid/liquid interfaces. It allows a wide range of possible investigations, including soft and biological samples under physiological conditions (hydrated specimens). The structural information obtained using the X-rays is combined with the data gathered with the AFM (morphology and mechanical properties), providing a unique characterization of the specimen and its dynamics in situ during an experiment. In this work, lipid monolayers and bilayers in air or liquid environment have been investigated by means of AFM, both with imaging and force spectroscopy, and X-ray reflectivity. In addition, this combination allows the radiation damage induced by the beam on the sample to be studied, as has been observed on DOPC and DPPC supported lipid bilayers under physiological conditions.

  8. AFM ve farmaceutické technologii 3.

    OpenAIRE

    Ščuryová, Veronika

    2015-01-01

    Charles University in Prague Faculty of Pharmacy in Hradec Králové Department of Pharmaceutical Technology Student: Veronika Ščuryová Supervisor: doc. RNDr. Pavel Doležal, CSc. Title of thesis: AFM in Pharmaceutical Technology 3 The theoretical part deals first with the construction of AFM microscope, the principle of the method, determining the surface topography and regimes which can be used. Described therein are distinct advantages over previous traditional methods but also its pitfalls. ...

  9. High-speed AFM of human chromosomes in liquid

    Energy Technology Data Exchange (ETDEWEB)

    Picco, L M; Dunton, P G; Ulcinas, A; Engledew, D J; Miles, M J [H H Wills Physics Laboratory and IRC in Nanotechnology, University of Bristol, Tyndall Avenue, Bristol BS8 1TL (United Kingdom); Hoshi, O; Ushiki, T [Division of Microscopic Anatomy and Bio-Imaging, Department of Cellular Function, Niigata University Graduate School of Medical and Dental Sciences, Asahimachi-Dori 1, Niigata, 951-8150 (Japan)], E-mail: m.j.miles@bristol.ac.uk

    2008-09-24

    Further developments of the previously reported high-speed contact-mode AFM are described. The technique is applied to the imaging of human chromosomes at video rate both in air and in water. These are the largest structures to have been imaged with high-speed AFM and the first imaging in liquid to be reported. A possible mechanism that allows such high-speed contact-mode imaging without significant damage to the sample is discussed in the context of the velocity dependence of the measured lateral force on the AFM tip.

  10. Surface features on Sahara soil dust particles made visible by atomic force microscope (AFM) phase images

    OpenAIRE

    G. Helas; M. O. Andreae

    2008-01-01

    We show that atomic force microscopy (AFM) phase images can reveal surface features of soil dust particles, which are not evident using other microscopic methods. The non-contact AFM method is able to resolve topographical structures in the nanometer range as well as to uncover repulsive atomic forces and attractive van der Waals' forces, and thus gives insight to surface properties. Though the method does not allow quantitative assignment in terms of chemical compound description, it clearly...

  11. Nanomechanical probing of soft matter through hydrophobic AFM tips fabricated by two-photon polymerization

    International Nuclear Information System (INIS)

    Suriano, Raffaella; De Marco, Carmela; Turri, Stefano; Zandrini, Tommaso; Osellame, Roberto; Bragheri, Francesca

    2016-01-01

    Atomic force microscopy (AFM) nanoindentation of soft materials is a powerful tool for probing mechanical properties of biomaterials. Though many results have been reported in this field over the last decade, adhesion forces between the tip and the sample hinder the elastic modulus measurement when hydrophilic soft samples are investigated. Here, two-photon polymerization (2PP) technology was used to fabricate hydrophobic perfluoropolyether-based AFM tips. The hydrophobic 2PP tips allowed us to overcome the limitations of commercial and functionalized tips as well as to successfully measure the elastic modulus of medically relevant soft materials in air. Our results obtained in the characterization of poly(dimethyl siloxane) and polyethylene glycol hydrogels showed lower adhesion forces over a larger measurement range when compared to measurements performed with commercial tips. The elastic moduli measured by means of hydrophobic 2PP AFM tips were also found to be comparable to those obtained using conventional techniques for macroscopic samples. We successfully showed that the hydrophobic AFM tips developed by this highly versatile technology enable the study of mechanical properties of soft matter, benefiting from reduced sample–tip interactions, and a custom-made shape and dimension of the tips. (paper)

  12. Attractive forces between hydrophobic solid surfaces measured by AFM on the first approach in salt solutions and in the presence of dissolved gases.

    Science.gov (United States)

    Azadi, Mehdi; Nguyen, Anh V; Yakubov, Gleb E

    2015-02-17

    Interfacial gas enrichment of dissolved gases (IGE) has been shown to cover hydrophobic solid surfaces in water. The atomic force microscopy (AFM) data has recently been supported by molecular dynamics simulation. It was demonstrated that IGE is responsible for the unexpected stability and large contact angle of gaseous nanobubbles at the hydrophobic solid-water interface. Here we provide further evidence of the significant effect of IGE on an attractive force between hydrophobic solid surfaces in water. The force in the presence of dissolved gas, i.e., in aerated and nonaerated NaCl solutions (up to 4 M), was measured by the AFM colloidal probe technique. The effect of nanobubble bridging on the attractive force was minimized or eliminated by measuring forces on the first approach of the AFM probe toward the flat hydrophobic surface and by using high salt concentrations to reduce gas solubility. Our results confirm the presence of three types of forces, two of which are long-range attractive forces of capillary bridging origin as caused by either surface nanobubbles or gap-induced cavitation. The third type is a short-range attractive force observed in the absence of interfacial nanobubbles that is attributed to the IGE in the form of a dense gas layer (DGL) at hydrophobic surfaces. Such a force was found to increase with increasing gas saturation and to decrease with decreasing gas solubility.

  13. Synthesis and AFM visualization of DNA nanostructures

    International Nuclear Information System (INIS)

    Mizuno, Rika; Haruta, Hirotaka; Morii, Takashi; Okada, Takao; Asakawa, Takeshi; Hayashi, Kenshi

    2004-01-01

    We propose a novel bottom-up approach for the fabrication of various desired nanostructures, based on self-assembly of oligonucleotides governed by Watson-Crick base pairing. Using this approach, we designed Y-shaped, closed Y-shaped, H-shaped, and hexagonal structures with oligonucleotides. These structures were autonomously fabricated simply by mixing equimolar solutions of oligonucleotides and performing hybridization. After synthesis of the nanostructures, we confirmed their validity by agarose gel electrophoresis and atomic force microscope (AFM) visualization. We detected bands of the desired molecular sizes in the gel electrophoresis and observed the desired structures by AFM analysis. We concluded that the synthesized structures were consistent with our intended design and that AFM visualization is a very useful tool for the observation of nanostructures

  14. Influence of smectite suspension structure on sheet orientation in dry sediments: XRD and AFM applications.

    Science.gov (United States)

    Zbik, Marek S; Frost, Ray L

    2010-06-15

    The structure-building phenomena within clay aggregates are governed by forces acting between clay particles. Measurements of such forces are important to understand in order to manipulate the aggregate structure for applications such as dewatering of mineral processing tailings. A parallel particle orientation is required when conducting XRD investigation on the oriented samples and conduct force measurements acting between basal planes of clay mineral platelets using atomic force microscopy (AFM). To investigate how smectite clay platelets were oriented on silicon wafer substrate when dried from suspension range of methods like SEM, XRD and AFM were employed. From these investigations, we conclude that high clay concentrations and larger particle diameters (up to 5 microm) in suspension result in random orientation of platelets in the substrate. The best possible laminar orientation in the clay dry film, represented in the XRD 001/020 intensity ratio of 47 was obtained by drying thin layers from 0.02 wt.% clay suspensions of the natural pH. Conducted AFM investigations show that smectite studied in water based electrolytes show very long-range repulsive forces lower in strength than electrostatic forces from double-layer repulsion. It was suggested that these forces may have structural nature. Smectite surface layers rehydrate in water environment forms surface gel with spongy and cellular texture which cushion approaching AFM probe. This structural effect can be measured in distances larger than 1000 nm from substrate surface and when probe penetrate this gel layer, structural linkages are forming between substrate and clay covered probe. These linkages prevent subsequently smooth detachments of AFM probe on way back when retrieval. This effect of tearing new formed structure apart involves larger adhesion-like forces measured in retrieval. It is also suggested that these effect may be enhanced by the nano-clay particles interaction. 2010 Elsevier Inc. All

  15. Hydration states of AFm cement phases

    Energy Technology Data Exchange (ETDEWEB)

    Baquerizo, Luis G., E-mail: luis.baquerizoibarra@holcim.com [Innovation, Holcim Technology Ltd., CH-5113 Holderbank (Switzerland); Matschei, Thomas [Innovation, Holcim Technology Ltd., CH-5113 Holderbank (Switzerland); Scrivener, Karen L. [Laboratory of Construction Materials, Ecole Polytechnique Fédérale de Lausanne, CH-1015 Lausanne (Switzerland); Saeidpour, Mahsa; Wadsö, Lars [Building Materials, Lund University, Box 124, 221 000 Lund (Sweden)

    2015-07-15

    The AFm phase, one of the main products formed during the hydration of Portland and calcium aluminate cement based systems, belongs to the layered double hydrate (LDH) family having positively charged layers and water plus charge-balancing anions in the interlayer. It is known that these phases present different hydration states (i.e. varying water content) depending on the relative humidity (RH), temperature and anion type, which might be linked to volume changes (swelling and shrinkage). Unfortunately the stability conditions of these phases are insufficiently reported. This paper presents novel experimental results on the different hydration states of the most important AFm phases: monocarboaluminate, hemicarboaluminate, strätlingite, hydroxy-AFm and monosulfoaluminate, and the thermodynamic properties associated with changes in their water content during absorption/desorption. This data opens the possibility to model the response of cementitious systems during drying and wetting and to engineer systems more resistant to harsh external conditions.

  16. Electrostatic nanopatterning of PMMA by AFM charge writing for directed nano-assembly

    International Nuclear Information System (INIS)

    Ressier, L; Nader, V Le

    2008-01-01

    Electrostatic nanopatterning of poly(methylmethacrylate) (PMMA) thin films by atomic force microscopy (AFM) charge writing was investigated using Kelvin force microscopy (KFM). The lateral size of the electrostatic patterns and the amount of injected charges are closely correlated and can be controlled by the height of the voltage pulses applied to the AFM tip and the tip-sample separation during the writing process. Charge retention measurements show that PMMA has excellent charge storage properties in air under relative humidities from 1% to 60% and withstands immersion in ultra-pure water. This study thus reveals that PMMA is a very promising electret to create efficient electrostatic nanopatterns for directed self-assembly of nanoscale objects, including the broad range of colloidal particles or molecules in aqueous solutions

  17. U.S. Army Training and Testing Area Carrying Capacity (ATTACC) for Munitions (AFM)

    Science.gov (United States)

    2006-11-01

    Army Training Support Center USDA United States Department of Agriculture USGS United States Geological Survey USLE Universal Soil Loss Equation...Range condition is a function of climate, soil , and hydrology. The munitions impact, constituent load, and range condition are modeled using AFM...For ArcGIS v2 to attain expected concentrations of munitions constituents and corresponding risk due to exposure through soil - and water-related

  18. AFM studies on heavy ion irradiated YBCO single crystals

    International Nuclear Information System (INIS)

    Lakhani, Archana; Marhas, M.K.; Saravanan, P.; Ganesan, V.; Srinivasan, R.; Kanjilal, D.; Mehta, G.K.; Elizabeth, Suja; Bhat, H.L.

    2000-01-01

    Atomic Force Microscopy (AFM) is extensively used to characterise the surface morphology of high energy ion irradiated single crystals of high temperature superconductor - YBCO. Our earlier systematic studies on thin films of YBCO under high energy and heavy ion irradiation shows clear evidence of ion induced sputtering or erosion, even though the effect is more on the grain boundaries. These earlier results were supported by electrical resistance measurements. In order to understand more clearly, the nature of surface modification at these high energies, AFM studies were carried out on single crystals of YBCO. Single crystals were chosen in order to see the effect on crystallites alone without interference from grain boundaries. 200 MeV gold ions were used for investigation using the facilities available at Nuclear Science Centre, New Delhi. The type of ion and the range of energies were chosen to meet the threshold for electronically mediated defect production. The results are in conformity with our earlier studies and will be described in detail in the context of electronic energy loss mediated sputtering or erosion. (author)

  19. Application of focused ion beam for the fabrication of AFM probes

    Science.gov (United States)

    Kolomiytsev, A. S.; Lisitsyn, S. A.; Smirnov, V. A.; Fedotov, A. A.; Varzarev, Yu N.

    2017-10-01

    The results of an experimental study of the probe tips fabrication for critical-dimension atomic force microscopy (CD-AFM) using the focused ion beam (FIB) induced deposition are presented. Methods of the FIB-induced deposition of tungsten and carbon onto the tip of an AFM probe are studied. Based on the results obtained in the study, probes for the CD-AFM technique with a tip height about 1 μm and radius of 20 nm were created. The formation of CD-AFM probes by FIB-induced deposition allows creating a high efficiency tool for nanotechnology and nanodiagnostics. The use of modified cantilevers allows minimizing the artefacts of AFM images and increasing the accuracy of the relief measurement. The obtained results can be used for fabrication of AFM probes for express monitoring of the technological process in the manufacturing of the elements for micro- and nanoelectronics.

  20. Improving the speed of AFM by mechatronic design and modern control methods

    International Nuclear Information System (INIS)

    Schitter, Georg

    2009-01-01

    In Atomic Force Microscopy (AFM) high-performance and high-precision control of the AFM scanner and of the imaging forces is crucial. Particularly at high imaging speeds the dynamic behaviour of the scanner may cause imaging artifacts and limit the maximum imaging rate. This contribution discusses and presents recent improvements in AFM instrumentation for faster imaging by means of mechatronic design and utilizing modern control engineering methods. Combining these improvements enables AFM imaging at more than two orders of magnitudes faster than conventional AFMs. (orig.)

  1. Atom-resolved AFM imaging of calcite nanoparticles in water

    Energy Technology Data Exchange (ETDEWEB)

    Imada, Hirotake; Kimura, Kenjiro [Department of Chemistry, School of Science, Kobe University, Rokko-dai, Nada, Kobe 657-8501 (Japan); Onishi, Hiroshi, E-mail: oni@kobe-u.ac.jp [Department of Chemistry, School of Science, Kobe University, Rokko-dai, Nada, Kobe 657-8501 (Japan)

    2013-06-20

    Highlights: ► An advanced frequency-modulation AFM (FM-AFM) was applied for imaging particles. ► Atom-resolved topography of nano-sized particles of calcite was observed in water. ► Locally ordered structures were found and assigned to a (104) facet of calcite. ► A promising ability of FM-AFM was demonstrated in imaging nano-sized particles. - Abstract: The atom-resolved topography of calcite nanoparticles was observed in water using a frequency-modulation atomic force microscope. Locally ordered structures were found and assigned to a (104) facet of crystalline calcite.

  2. Atom-resolved AFM imaging of calcite nanoparticles in water

    International Nuclear Information System (INIS)

    Imada, Hirotake; Kimura, Kenjiro; Onishi, Hiroshi

    2013-01-01

    Highlights: ► An advanced frequency-modulation AFM (FM-AFM) was applied for imaging particles. ► Atom-resolved topography of nano-sized particles of calcite was observed in water. ► Locally ordered structures were found and assigned to a (104) facet of calcite. ► A promising ability of FM-AFM was demonstrated in imaging nano-sized particles. - Abstract: The atom-resolved topography of calcite nanoparticles was observed in water using a frequency-modulation atomic force microscope. Locally ordered structures were found and assigned to a (104) facet of crystalline calcite

  3. Analysis of aflatoxin M1 in milk and yogurt and AFM1 reduction by lactic acid bacteria used in Lebanese industry

    International Nuclear Information System (INIS)

    El Khoury, A.; Yaghi, J.; Atoui, A.

    2011-01-01

    In this study a screening survey was undertaken to determine the presence and levels of aflatoxin M1 (AFM1) in locally produced dairy products. For this purpose, a total of 138 dairy samples (milk and yogurt) were analyzed to quantify AFM1 levels in these products. Results obtained showed that AFM1 was found in 40.62% and 32.81% of milk and yogurt samples respectively. The range of contamination levels varied between lower and higher than European regulation limit. Lactic acid bacteria (Lactobacillus bulgaricus and Streptococcus thermophilus) used in the Lebanese traditional industry were screened for their ability to reduce the level of AFM1. Due to the lack in data on the natural occurrence of AFM1 in Lebanese dairy products, the aim of this work was to report some information and to state these products according to EC regulations. (author)

  4. High-stress study of bioinspired multifunctional PEDOT:PSS/nanoclay nanocomposites using AFM, SEM and numerical simulation

    Directory of Open Access Journals (Sweden)

    Alfredo J. Diaz

    2017-10-01

    Full Text Available Bioinspired design has been central in the development of hierarchical nanocomposites. Particularly, the nacre-mimetic brick-and-mortar structure has shown excellent mechanical properties, as well as gas-barrier properties and optical transparency. Along with these intrinsic properties, the layered structure has also been utilized in sensing devices. Here we extend the multifunctionality of nacre-mimetics by designing an optically transparent and electron conductive coating based on PEDOT:PSS and nanoclays Laponite RD and Cloisite Na+. We carry out extensive characterization of the nanocomposite using transmittance spectra (transparency, conductive atomic force microscopy (conductivity, contact-resonance force microscopy (mechanical properties, and SEM combined with a variety of stress-strain AFM experiments and AFM numerical simulations (internal structure. We further study the nanoclay’s response to the application of pressure with multifrequency AFM and conductive AFM, whereby increases and decreases in conductivity can occur for the Laponite RD composites. We offer a possible mechanism to explain the changes in conductivity by modeling the coating as a 1-dimensional multibarrier potential for electron transport, and show that conductivity can change when the separation between the barriers changes under the application of pressure, and that the direction of the change depends on the energy of the electrons. We did not observe changes in conductivity under the application of pressure with AFM for the Cloisite Na+ nanocomposite, which has a large platelet size compared with the AFM probe diameter. No pressure-induced changes in conductivity were observed in the clay-free polymer either.

  5. Sequential electrochemical oxidation and site-selective growth of nanoparticles onto AFM probes.

    Science.gov (United States)

    Wang, Haitao; Tian, Tian; Zhang, Yong; Pan, Zhiqiang; Wang, Yong; Xiao, Zhongdang

    2008-08-19

    In this work, we reported an approach for the site-selective growth of nanoparticle onto the tip apex of an atomic force microscopy (AFM) probe. The silicon AFM probe was first coated with a self-assembled monolayer (SAM) of octadecyltrichlorosilane (OTS) through a chemical vapor deposition (CVD) method. Subsequently, COOH groups were selectively generated at the tip apex of silicon AFM probes by applying an appropriate bias voltage between the tip and a flat gold electrode. The transformation of methyl to carboxylic groups at the tip apex of the AFM probe was investigated through measuring the capillary force before and after electrochemical oxidation. To prepare the nanoparticle terminated AFM probe, the oxidized AFM probe was then immersed in an aqueous solution containing positive metal ions, for example, Ag+, to bind positive metal ions to the oxidized area (COOH terminated area), followed by chemical reduction with aqueous NaBH 4 and further development (if desired) to give a metal nanoparticle-modified AFM probe. The formation of a metal nanoparticle at the tip apex of the AFM probe was confirmed by scanning electron microscopy (SEM) and energy dispersive X-ray analysis (EDXA).

  6. The spin-3/2 Ising model AFM/AFM two-layer lattice with crystal field

    International Nuclear Information System (INIS)

    Yigit, A.; Albayrak, E.

    2010-01-01

    The spin-3/2 Ising model is investigated for the case of antiferromagnetic (AFM/AFM) interactions on the two-layer Bethe lattice by using the exact recursion relations in a pairwise approach for given coordination numbers q=3, 4 and 6 when the layers are under the influences of equal external magnetic and equal crystal fields. The ground state (GS) phase diagrams are obtained on the different planes in detail and then the temperature dependent phase diagrams of the system are calculated accordingly. It is observed that the system presents both second- and first-order phase transitions for all q, therefore, tricritical points. It was also found that the system exhibits double-critical end points and isolated points. The model also presents two Neel temperatures, TN, and the existence of which leads to the reentrant behavior.

  7. Recent developments in dimensional nanometrology using AFMs

    Science.gov (United States)

    Yacoot, Andrew; Koenders, Ludger

    2011-12-01

    Scanning probe microscopes, in particular the atomic force microscope (AFM), have developed into sophisticated instruments that, throughout the world, are no longer used just for imaging, but for quantitative measurements. A role of the national measurement institutes has been to provide traceable metrology for these instruments. This paper presents a brief overview as to how this has been achieved, highlights the future requirements for metrology to support developments in AFM technology and describes work in progress to meet this need.

  8. Recent developments in dimensional nanometrology using AFMs

    International Nuclear Information System (INIS)

    Yacoot, Andrew; Koenders, Ludger

    2011-01-01

    Scanning probe microscopes, in particular the atomic force microscope (AFM), have developed into sophisticated instruments that, throughout the world, are no longer used just for imaging, but for quantitative measurements. A role of the national measurement institutes has been to provide traceable metrology for these instruments. This paper presents a brief overview as to how this has been achieved, highlights the future requirements for metrology to support developments in AFM technology and describes work in progress to meet this need. (perspective)

  9. Multiple regimes of operation in bimodal AFM: understanding the energy of cantilever eigenmodes

    Directory of Open Access Journals (Sweden)

    Daniel Kiracofe

    2013-06-01

    Full Text Available One of the key goals in atomic force microscopy (AFM imaging is to enhance material property contrast with high resolution. Bimodal AFM, where two eigenmodes are simultaneously excited, confers significant advantages over conventional single-frequency tapping mode AFM due to its ability to provide contrast between regions with different material properties under gentle imaging conditions. Bimodal AFM traditionally uses the first two eigenmodes of the AFM cantilever. In this work, the authors explore the use of higher eigenmodes in bimodal AFM (e.g., exciting the first and fourth eigenmodes. It is found that such operation leads to interesting contrast reversals compared to traditional bimodal AFM. A series of experiments and numerical simulations shows that the primary cause of the contrast reversals is not the choice of eigenmode itself (e.g., second versus fourth, but rather the relative kinetic energy between the higher eigenmode and the first eigenmode. This leads to the identification of three distinct imaging regimes in bimodal AFM. This result, which is applicable even to traditional bimodal AFM, should allow researchers to choose cantilever and operating parameters in a more rational manner in order to optimize resolution and contrast during nanoscale imaging of materials.

  10. AFM indentation study of breast cancer cells

    International Nuclear Information System (INIS)

    Li, Q.S.; Lee, G.Y.H.; Ong, C.N.; Lim, C.T.

    2008-01-01

    Mechanical properties of individual living cells are known to be closely related to the health and function of the human body. Here, atomic force microscopy (AFM) indentation using a micro-sized spherical probe was carried out to characterize the elasticity of benign (MCF-10A) and cancerous (MCF-7) human breast epithelial cells. AFM imaging and confocal fluorescence imaging were also used to investigate their corresponding sub-membrane cytoskeletal structures. Malignant (MCF-7) breast cells were found to have an apparent Young's modulus significantly lower (1.4-1.8 times) than that of their non-malignant (MCF-10A) counterparts at physiological temperature (37 deg. C), and their apparent Young's modulus increase with loading rate. Both confocal and AFM images showed a significant difference in the organization of their sub-membrane actin structures which directly contribute to their difference in cell elasticity. This change may have facilitated easy migration and invasion of malignant cells during metastasis

  11. AFM tip-sample convolution effects for cylinder protrusions

    Science.gov (United States)

    Shen, Jian; Zhang, Dan; Zhang, Fei-Hu; Gan, Yang

    2017-11-01

    A thorough understanding about the AFM tip geometry dependent artifacts and tip-sample convolution effect is essential for reliable AFM topographic characterization and dimensional metrology. Using rigid sapphire cylinder protrusions (diameter: 2.25 μm, height: 575 nm) as the model system, a systematic and quantitative study about the imaging artifacts of four types of tips-two different pyramidal tips, one tetrahedral tip and one super sharp whisker tip-is carried out through comparing tip geometry dependent variations in AFM topography of cylinders and constructing the rigid tip-cylinder convolution models. We found that the imaging artifacts and the tip-sample convolution effect are critically related to the actual inclination of the working cantilever, the tip geometry, and the obstructive contacts between the working tip's planes/edges and the cylinder. Artifact-free images can only be obtained provided that all planes and edges of the working tip are steeper than the cylinder sidewalls. The findings reported here will contribute to reliable AFM characterization of surface features of micron or hundreds of nanometers in height that are frequently met in semiconductor, biology and materials fields.

  12. Magnetotransport measurements on AFM structured two-dimensional electron gases on cleaved edges of GaAs/AlGaAs; Magnetotransportmessungen an AFM-strukturierten zweidimensionalen Elektronengasen auf GaAs/AlGaAs-Spaltkanten

    Energy Technology Data Exchange (ETDEWEB)

    Reinwald, Elisabeth

    2009-06-25

    In this thesis a two dimensional electron gas (2DEG) on a (110) cleavage plane of a GaAs/AlGaAs(001) heterostructure was produced by means of cleaved edge overgrowth (CEO) and modulated in two dimensions. The 2DEG was modulated in one direction by a superlattice of the subjacent GaAs/AlGaAs(001) heterostructure. A second modulation, perpendicular to the first was realized by local anodic oxidation (LAO) with an atomic force microscope (AFM). For the process of LAO an electric voltage is applied between the tip of the AFM and the surface of the GaAs. The natural water film on the surface acts as electrolyte so that the GaAs surface is locally oxidized underneath the AFM tip. This oxide leads to a band bending so that the 2DEG underneath the oxide is locally depleted. On these systems magnetotransport measurements revealed that it is actually possible to modulate 2DEGs on a sufficient large area by local anodic oxidation. On the cleaved surfaces the influence of the two dimensional modulation on the electron gas has been demonstrated. (orig.)

  13. Direct visualization of glutamate transporter elevator mechanism by high-speed AFM.

    Science.gov (United States)

    Ruan, Yi; Miyagi, Atsushi; Wang, Xiaoyu; Chami, Mohamed; Boudker, Olga; Scheuring, Simon

    2017-02-14

    Glutamate transporters are essential for recovery of the neurotransmitter glutamate from the synaptic cleft. Crystal structures in the outward- and inward-facing conformations of a glutamate transporter homolog from archaebacterium Pyrococcus horikoshii , sodium/aspartate symporter Glt Ph , suggested the molecular basis of the transporter cycle. However, dynamic studies of the transport mechanism have been sparse and indirect. Here we present high-speed atomic force microscopy (HS-AFM) observations of membrane-reconstituted Glt Ph at work. HS-AFM movies provide unprecedented real-space and real-time visualization of the transport dynamics. Our results show transport mediated by large amplitude 1.85-nm "elevator" movements of the transport domains consistent with previous crystallographic and spectroscopic studies. Elevator dynamics occur in the absence and presence of sodium ions and aspartate, but stall in sodium alone, providing a direct visualization of the ion and substrate symport mechanism. We show unambiguously that individual protomers within the trimeric transporter function fully independently.

  14. AFM analysis of bleaching effects on dental enamel microtopography

    Energy Technology Data Exchange (ETDEWEB)

    Pedreira de Freitas, Ana Carolina, E-mail: anacarolfreitas@usp.br [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Cardoso Espejo, Luciana, E-mail: luespejo@hotmail.com [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Brossi Botta, Sergio, E-mail: sbbotta@usp.br [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Sa Teixeira, Fernanda de, E-mail: nandast@if.usp.br [Laboratorio de Filmes Finos, Instituto de Fisica da Universidade de Sao Paulo, Rua do Matao, Travessa R, 187 - Cidade Universitaria, CEP 05314-970, Sao Paulo (Brazil); Cerqueira, Luz Maria Aparecida A., E-mail: maacluz@usp.br [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Garone-Netto, Narciso, E-mail: ngarone@usp.br [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Bona Matos, Adriana, E-mail: bona@usp.br [Departamento de Dentistica, Faculdade de Odontologia da Universidade de Sao Paulo, Av. Prof. Lineu Prestes, 2227 - Cidade Universitaria, CEP 05508-000, Sao Paulo (Brazil); Barbosa da Silveira Salvadori, Maria Cecilia, E-mail: mcsalva@if.usp.br [Laboratorio de Filmes Finos, Instituto de Fisica da Universidade de Sao Paulo, Rua do Matao, Travessa R, 187 - Cidade Universitaria, CEP 05314-970, Sao Paulo (Brazil)

    2010-02-15

    The purpose of this in vitro study was to test a new methodology to evaluate the effects of 35% hydrogen peroxide agent on the microtopography of sound enamel using an atomic force microscope (AFM). The buccal sound surfaces of three extracted human lower incisors were used, without polishing the surfaces to maintain them with natural morphology. These unpolished surfaces were subjected to bleaching procedure with 35% hydrogen peroxide that consisted of 4 applications of the bleaching agent on enamel surfaces for 10 min each application. Surface images were obtained in a 15 {mu}m x 15 {mu}m area using an AFM. The roughness (Ra and RMS) and the power spectral density (PSD) were obtained before and after the bleaching treatment. As results we could inquire that the PSD analyses were very suitable to identifying the morphological changes on the surfaces, while the Ra and RMS parameters were insufficient to represent the morphological alterations promoted by bleaching procedure on enamel. The morphological wavelength in the range of visible light spectrum (380-750 nm) was analyzed, showing a considerable increase of the PSD with the bleaching treatment.

  15. AFM analysis of bleaching effects on dental enamel microtopography

    International Nuclear Information System (INIS)

    Pedreira de Freitas, Ana Carolina; Cardoso Espejo, Luciana; Brossi Botta, Sergio; Sa Teixeira, Fernanda de; Cerqueira, Luz Maria Aparecida A.; Garone-Netto, Narciso; Bona Matos, Adriana; Barbosa da Silveira Salvadori, Maria Cecilia

    2010-01-01

    The purpose of this in vitro study was to test a new methodology to evaluate the effects of 35% hydrogen peroxide agent on the microtopography of sound enamel using an atomic force microscope (AFM). The buccal sound surfaces of three extracted human lower incisors were used, without polishing the surfaces to maintain them with natural morphology. These unpolished surfaces were subjected to bleaching procedure with 35% hydrogen peroxide that consisted of 4 applications of the bleaching agent on enamel surfaces for 10 min each application. Surface images were obtained in a 15 μm x 15 μm area using an AFM. The roughness (Ra and RMS) and the power spectral density (PSD) were obtained before and after the bleaching treatment. As results we could inquire that the PSD analyses were very suitable to identifying the morphological changes on the surfaces, while the Ra and RMS parameters were insufficient to represent the morphological alterations promoted by bleaching procedure on enamel. The morphological wavelength in the range of visible light spectrum (380-750 nm) was analyzed, showing a considerable increase of the PSD with the bleaching treatment.

  16. SEM and AFM studies of dip-coated CuO nanofilms.

    Science.gov (United States)

    Dhanasekaran, V; Mahalingam, T; Ganesan, V

    2013-01-01

    Cupric oxide (CuO) semiconducting thin films were prepared at various copper sulfate concentrations by dip coating. The copper sulfate concentration was varied to yield films of thicknesses in the range of 445-685 nm by surface profilometer. X-ray diffraction patterns revealed that the deposited films were polycrystalline in nature with monoclinic structure of (-111) plane. The surface morphology and topography of monoclinic-phase CuO thin films were examined using scanning electron microscopy (SEM) and atomic force microscopy (AFM), respectively. Surface roughness profile was plotted using WSxM software and the estimated surface roughness was about ∼19.4 nm at 30 mM molar concentration. The nanosheets shaped grains were observed by SEM and AFM studies. The stoichiometric compound formation was observed at 30 mM copper sulfate concentration prepared film by EDX. The indirect band gap energy of CuO films was increased from 1.08 to 1.20 eV with the increase of copper sulfate concentrations. Copyright © 2012 Wiley Periodicals, Inc.

  17. Combining adhesive contact mechanics with a viscoelastic material model to probe local material properties by AFM.

    Science.gov (United States)

    Ganser, Christian; Czibula, Caterina; Tscharnuter, Daniel; Schöberl, Thomas; Teichert, Christian; Hirn, Ulrich

    2017-12-20

    Viscoelastic properties are often measured using probe based techniques such as nanoindentation (NI) and atomic force microscopy (AFM). Rarely, however, are these methods verified. In this article, we present a method that combines contact mechanics with a viscoelastic model (VEM) composed of springs and dashpots. We further show how to use this model to determine viscoelastic properties from creep curves recorded by a probe based technique. We focus on using the standard linear solid model and the generalized Maxwell model of order 2. The method operates in the range of 0.01 Hz to 1 Hz. Our approach is suitable for rough surfaces by providing a defined contact area using plastic pre-deformation of the material. The very same procedure is used to evaluate AFM based measurements as well as NI measurements performed on polymer samples made from poly(methyl methacrylate) and polycarbonate. The results of these measurements are then compared to those obtained by tensile creep tests also performed on the same samples. It is found that the tensile test results differ considerably from the results obtained by AFM and NI methods. The similarity between the AFM results and NI results suggests that the proposed method is capable of yielding results comparable to NI but with the advantage of the imaging possibilities of AFM. Furthermore, all three methods allowed a clear distinction between PC and PMMA by means of their respective viscoelastic properties.

  18. AFM (Atomic force microscope and its use in studying the surface

    Directory of Open Access Journals (Sweden)

    Škvarla Jiří

    1996-06-01

    Full Text Available The paper summarizes the present knowledge about the use of AFM in the mineral processing research. First, the development and fundamentals of the AFM imaging are presented in relation to other imaging techniques (especially STM, Scanning tunneling microscope. Further, the role of the sensing tip-surface interactions is mentioned. Finally, the surface force measurements in the AFM force calibration mode are diskussed.

  19. Helium ion beam induced growth of hammerhead AFM probes

    NARCIS (Netherlands)

    Nanda, G.; Veldhoven, E. van; Maas, D.J.; Sadeghian Marnani, H.; Alkemade, P.F.A.

    2015-01-01

    The authors report the direct-write growth of hammerhead atomic force microscope (AFM) probes by He+ beam induced deposition of platinum-carbon. In order to grow a thin nanoneedle on top of a conventional AFM probe, the authors move a focused He+ beam during exposure to a PtC precursor gas. In the

  20. Immobilisation of living bacteria for AFM imaging under physiological conditions

    International Nuclear Information System (INIS)

    Louise Meyer, Rikke; Zhou, Xingfei; Tang, Lone; Arpanaei, Ayyoob; Kingshott, Peter; Besenbacher, Flemming

    2010-01-01

    Atomic force microscopy (AFM) holds great potential for studying the nanoscale surface structures of living cells, and to measure their interactions with abiotic surfaces, other cells, or specific biomolecules. However, the application of AFM in microbiology is challenging due to the difficulty of immobilising bacterial cells to a flat surface without changing the cell surface properties or cell viability. We have performed an extensive and thorough study of how to functionalise surfaces in order to immobilise living bacteria for AFM studies in liquid environments. Our aim was to develop a scheme which allows bacterial cells to be immobilised to a flat surface with sufficient strength to avoid detachment during the AFM scanning, and without affecting cell surface chemistry, structure, and viability. We compare and evaluate published methods, and present a new, reproducible, and generally applicable scheme for immobilising bacteria cells for an AFM imaging. Bacterial cells were immobilised to modified glass surfaces by physical confinement of cells in microwells, physisorption to positively charged surfaces, covalent binding to amine- or carboxyl-terminated surfaces, and adsorption to surfaces coated with highly adhesive polyphenolic proteins originating from the mussel Mytilus edulis. Living cells could be immobilised with all of these approaches, but many cells detached when immobilised by electrostatic interactions and imaged in buffers like PBS or MOPS. Cells were more firmly attached when immobilised by covalent binding, although some cells still detached during AFM imaging. The most successful method revealed was immobilisation by polyphenolic proteins, which facilitated firm immobilisation of the cells. Furthermore, the cell viability was not affected by this immobilisation scheme, and adhesive proteins thus provide a fast, reproducible, and generally applicable scheme for immobilising living bacteria for an AFM imaging.

  1. Multifractural analysis of AFM images of Nb thin film surfaces

    International Nuclear Information System (INIS)

    Altajskij, M.V; Chernenko, L.P.; Balebanov, V.M.; Erokhin, N.S.; Moiseev, S.S.

    2000-01-01

    The multifractal analysis of the atomic Force Microscope (AFM) images of the Niobium (Nb) thin film surfaces has been performed. These Nb films are being used for the measurements of the London penetration depth of stationary magnetic field by polarized neutron reflectometry. The analysis shows the behavior of Renyi dimensions of images (in the range of available scales 6-2000 nm), like the known multifractal p-model, with typical Hausdorff dimension of prevalent color in the range of 1.6-1.9. This indicates the fractal nature of film landscape on those scales. The perspective of new mechanism of order parameter suppression on superconductor-vacuum boundary, manifested in anomalous magnetic field penetration in discussed

  2. AFM lithography of aluminum for fabrication of nanomechanical systems

    DEFF Research Database (Denmark)

    Davis, Zachary James; Abadal, G.; Hansen, Ole

    2003-01-01

    Nanolithography by local anodic oxidation of surfaces using atomic force microscopy (AFM) has proven to be more reproducible when using dynamic, non-contact mode. Hereby, the tip/sample interaction forces are reduced dramatically compared to contact mode, and thus tip wear is greatly reduced....... Anodic oxidation of Al can be used for fabricating nanomechanical systems, by using the Al oxide as a highly selective dry etching mask. In our experiments, areas as large as 2 mum x 3 mum have been oxidized repeatedly without any sign of tip-wear. Furthermore, line widths down to 10 nm have been...

  3. Metrological large range scanning probe microscope

    International Nuclear Information System (INIS)

    Dai Gaoliang; Pohlenz, Frank; Danzebrink, Hans-Ulrich; Xu Min; Hasche, Klaus; Wilkening, Guenter

    2004-01-01

    We describe a metrological large range scanning probe microscope (LR-SPM) with an Abbe error free design and direct interferometric position measurement capability, aimed at versatile traceable topographic measurements that require nanometer accuracy. A dual-stage positioning system was designed to achieve both a large measurement range and a high measurement speed. This dual-stage system consists of a commercially available stage, referred to as nanomeasuring machine (NMM), with a motion range of 25 mmx25 mmx5 mm along x, y, and z axes, and a compact z-axis piezoelectric positioning stage (compact z stage) with an extension range of 2 μm. The metrological LR-SPM described here senses the surface using a stationary fixed scanning force microscope (SFM) head working in contact mode. During operation, lateral scanning of the sample is performed solely by the NMM. Whereas the z motion, controlled by the SFM signal, is carried out by a combination of the NMM and the compact z stage. In this case the compact z stage, with its high mechanical resonance frequency (greater than 20 kHz), is responsible for the rapid motion while the NMM simultaneously makes slower movements over a larger motion range. To reduce the Abbe offset to a minimum the SFM tip is located at the intersection of three interferometer measurement beams orientated in x, y, and z directions. To improve real time performance two high-end digital signal processing (DSP) systems are used for NMM positioning and SFM servocontrol. Comprehensive DSP firmware and Windows XP-based software are implemented, providing a flexible and user-friendly interface. The instrument is able to perform large area imaging or profile scanning directly without stitching small scanned images. Several measurements on different samples such as flatness standards, nanostep height standards, roughness standards as well as sharp nanoedge samples and 1D gratings demonstrate the outstanding metrological capabilities of the instrument

  4. Adhesion forces in AFM of redox responsive polymer grafts: Effects of tip hydrophilicity

    NARCIS (Netherlands)

    Feng, Xueling; Kieviet, B.D.; Song, Jing; Schön, Peter Manfred; Vancso, Gyula J.

    2014-01-01

    The adherence between silicon nitride AFM tips and redox-active poly(ferrocenylsilanes) (PFS) grafts ongold was investigated by electrochemical AFM force spectroscopy. Before the adhesion measurementssilicon nitride AFM probes were cleaned with organic solvents (acetone and ethanol) or piranha

  5. Micromechanical Characterization of Complex Polypropylene Morphologies by HarmoniX AFM

    Directory of Open Access Journals (Sweden)

    S. Liparoti

    2017-01-01

    Full Text Available This paper examines the capability of the HarmoniX Atomic Force Microscopy (AFM technique to draw accurate and reliable micromechanical characterization of complex polymer morphologies generally found in conventional thermoplastic polymers. To that purpose, injection molded polypropylene samples, containing representative morphologies, have been characterized by HarmoniX AFM. Mapping and distributions of mechanical properties of the samples surface are determined and analyzed. Effects of sample preparation and test conditions are also analyzed. Finally, the AFM determination of surface elastic moduli has been compared with that obtained by indentation tests, finding good agreement among the results.

  6. Noise in NC-AFM measurements with significant tip–sample interaction

    Directory of Open Access Journals (Sweden)

    Jannis Lübbe

    2016-12-01

    Full Text Available The frequency shift noise in non-contact atomic force microscopy (NC-AFM imaging and spectroscopy consists of thermal noise and detection system noise with an additional contribution from amplitude noise if there are significant tip–sample interactions. The total noise power spectral density DΔf(fm is, however, not just the sum of these noise contributions. Instead its magnitude and spectral characteristics are determined by the strongly non-linear tip–sample interaction, by the coupling between the amplitude and tip–sample distance control loops of the NC-AFM system as well as by the characteristics of the phase locked loop (PLL detector used for frequency demodulation. Here, we measure DΔf(fm for various NC-AFM parameter settings representing realistic measurement conditions and compare experimental data to simulations based on a model of the NC-AFM system that includes the tip–sample interaction. The good agreement between predicted and measured noise spectra confirms that the model covers the relevant noise contributions and interactions. Results yield a general understanding of noise generation and propagation in the NC-AFM and provide a quantitative prediction of noise for given experimental parameters. We derive strategies for noise-optimised imaging and spectroscopy and outline a full optimisation procedure for the instrumentation and control loops.

  7. Direct manipulation of intracellular stress fibres using a hook-shaped AFM probe

    International Nuclear Information System (INIS)

    Machida, Shinichi; Watanabe-Nakayama, Takahiro; Harada, Ichiro; Afrin, Rehana; Nakayama, Tomonobu; Ikai, Atsushi

    2010-01-01

    Atomic force microscopy (AFM) is a highly successful technique for imaging nanometre-sized samples and measuring pico- to nano-newton forces acting between atoms and molecules. When it comes to the manipulation of larger samples with forces of tens and hundreds of nano-newtons, however, the present chemistry-based modification protocols for functionalizing AFM cantilevers to achieve the formation of covalent/non-covalent linkages between the AFM probe and the sample surface do not produce strong enough bonds. For the purpose of measuring the fracture strength and other mechanical properties of stress fibres (SFs) in living as well as semi-intact fibroblast cells, we fabricated an AFM probe with a hooking function by focused ion beam technology and used the AFM probe hook to capture, pull and eventually sever a chosen SF labelled with green or red fluorescent protein.

  8. Surface features on Sahara soil dust particles made visible by atomic force microscope (AFM phase images

    Directory of Open Access Journals (Sweden)

    M. O. Andreae

    2008-10-01

    Full Text Available We show that atomic force microscopy (AFM phase images can reveal surface features of soil dust particles, which are not evident using other microscopic methods. The non-contact AFM method is able to resolve topographical structures in the nanometer range as well as to uncover repulsive atomic forces and attractive van der Waals' forces, and thus gives insight to surface properties. Though the method does not allow quantitative assignment in terms of chemical compound description, it clearly shows deposits of distinguishable material on the surface. We apply this technique to dust aerosol particles from the Sahara collected over the Atlantic Ocean and describe micro-features on the surfaces of such particles.

  9. Molecular dynamics simulation of AFM studies of a single polymer chain

    International Nuclear Information System (INIS)

    Wang Wenhai; Kistler, Kurt A.; Sadeghipour, Keya; Baran, George

    2008-01-01

    Single polymer chain force-extension behavior measured by Atomic Force Microscopy (AFM) was interpreted by molecular dynamics (MD) simulation performed by applying a bead-spring (coarse-graining) model in which the bond potential function between adjacent beads is described by a worm-like chain (WLC) model. Simulation results indicate that caution should be applied when interpreting experimental AFM data, because the data vary depending on the point of AFM tip-polymer chain attachment. This approach offers an effective way for eventual analysis of the mechanical behavior of complex polymer networks

  10. Molecular dynamics simulation of AFM studies of a single polymer chain

    Energy Technology Data Exchange (ETDEWEB)

    Wang Wenhai [Center for Bioengineering and Biomaterials, College of Engineering, Temple University, 1947 N. 12th Street, Philadelphia, PA 19122 (United States); Kistler, Kurt A. [Department of Chemistry, Temple University, 1901 N. 13th Street, Philadelphia, PA 19122 (United States); Sadeghipour, Keya [Center for Bioengineering and Biomaterials, College of Engineering, Temple University, 1947 N. 12th Street, Philadelphia, PA 19122 (United States); Baran, George [Center for Bioengineering and Biomaterials, College of Engineering, Temple University, 1947 N. 12th Street, Philadelphia, PA 19122 (United States)], E-mail: grbaran@temple.edu

    2008-11-24

    Single polymer chain force-extension behavior measured by Atomic Force Microscopy (AFM) was interpreted by molecular dynamics (MD) simulation performed by applying a bead-spring (coarse-graining) model in which the bond potential function between adjacent beads is described by a worm-like chain (WLC) model. Simulation results indicate that caution should be applied when interpreting experimental AFM data, because the data vary depending on the point of AFM tip-polymer chain attachment. This approach offers an effective way for eventual analysis of the mechanical behavior of complex polymer networks.

  11. A low-cost AFM setup with an interferometer for undergraduates and secondary-school students

    International Nuclear Information System (INIS)

    Bergmann, Antje; Feigl, Daniela; Kuhn, David; Schaupp, Manuel; Quast, Günter; Busch, Kurt; Eichner, Ludwig; Schumacher, Jens

    2013-01-01

    Atomic force microscopy (AFM) is an important tool in nanotechnology. This method makes it possible to observe nanoscopic surfaces beyond the resolution of light microscopy. In order to provide undergraduate and secondary-school students with insights into this world, we have developed a very robust low-cost AFM setup with a Fabry–Perot interferometer as a detecting device. This setup is designed to be operated almost completely manually and its simplicity gives access to a profound understanding of the working principle. Our AFM is operated in a constant height mode, i.e. the topography of the sample surface is represented directly by the deflection of the cantilever. Thus, the measuring procedure can be understood even by secondary-school students; furthermore, it is the method with the lowest cost, totalling not more than 10–15 k Euros. Nevertheless, we are able to examine a large variety of sample topographies such as CD and DVD surfaces, IC structures, blood cells, butterfly wings or moth eyes. Furthermore, force–distance curves can be recorded and the tensile moduli of some materials can be evaluated. We present our setup in detail and describe its working principles. In addition, we show various experiments which have already been performed by students. (paper)

  12. Modification of AFM Tips for Facilitating Picking-up of Nanoparticles

    International Nuclear Information System (INIS)

    Peng, Wang; Hai-Jun, Yang; Hua-Bin, Wang; Hai, Li; Xin-Yan, Wang; Ying, Wang; Jun-Hong, Lü; Bin, Li; Yi, Zhang; Jun, Hu

    2008-01-01

    The radius of atomic force microscope (AFM) tip is a key factor that influences nonspecific interactions between AFM tip and nanoparticles. Generally, a tip with larger radius contributes to a higher efficiency of picking up nanoparticles. We provide two methods for modifying the AFM tip: one is to wear a tip apex on a solid substrate and the other is to coat a tip with poly (dimethylsiloxane) (PDMS). Both the approaches can enhance the adhesion force between the tip and nanoparticles by increasing tip radius. The experimental results show that a modified tip, compared to an unmodified one, achieves six-fold efficiency improvement in the capture of targeted colloidal gold nanoparticles. (general)

  13. A Multifunctional Frontloading Approach for Repeated Recycling of a Pressure-Controlled AFM Micropipette.

    Directory of Open Access Journals (Sweden)

    Phillip Roder

    Full Text Available Fluid force microscopy combines the positional accuracy and force sensitivity of an atomic force microscope (AFM with nanofluidics via a microchanneled cantilever. However, adequate loading and cleaning procedures for such AFM micropipettes are required for various application situations. Here, a new frontloading procedure is described for an AFM micropipette functioning as a force- and pressure-controlled microscale liquid dispenser. This frontloading procedure seems especially attractive when using target substances featuring high costs or low available amounts. Here, the AFM micropipette could be filled from the tip side with liquid from a previously applied droplet with a volume of only a few μL using a short low-pressure pulse. The liquid-loaded AFM micropipettes could be then applied for experiments in air or liquid environments. AFM micropipette frontloading was evaluated with the well-known organic fluorescent dye rhodamine 6G and the AlexaFluor647-labeled antibody goat anti-rat IgG as an example of a larger biological compound. After micropipette usage, specific cleaning procedures were tested. Furthermore, a storage method is described, at which the AFM micropipettes could be stored for a few hours up to several days without drying out or clogging of the microchannel. In summary, the rapid, versatile and cost-efficient frontloading and cleaning procedure for the repeated usage of a single AFM micropipette is beneficial for various application situations from specific surface modifications through to local manipulation of living cells, and provides a simplified and faster handling for already known experiments with fluid force microscopy.

  14. Probing colloidal forces between a Si3N4 AFM tip and single nanoparticles of silica and alumina.

    Science.gov (United States)

    Drelich, J; Long, J; Xu, Z; Masliyah, J; White, C L

    2006-11-15

    The atomic force microscope (AFM) has been used to measure surface forces between silicon nitride AFM tips and individual nanoparticles deposited on substrates in 10(-4) and 10(-2) M KCl solutions. Silica nanoparticles (10 nm diameter) were deposited on an alumina substrate and alumina particles (5 to 80 nm diameter) were deposited on a mica substrate using aqueous suspensions. Ionic concentrations and pH were used to manage attractive substrate-particle electrostatic forces. The AFM tip was located on deposited nanoparticles using an operator controlled offset to achieve stepwise tip movements. Nanoparticles were found to have a negligible effect on long-range tip-substrate interactions, however, the forces between the tip and nanoparticle were detectable at small separations. Exponentially increasing short-range repulsive forces, attributed to the hydration forces, were observed for silica nanoparticles. The effective range of hydration forces was found to be 2-3 nm with the decay length of 0.8-1.3 nm. These parameters are in a good agreement with the results reported for macroscopic surfaces of silica obtained using the surface force apparatus suggesting that hydration forces for the silica nanoparticles are similar to those for flat silica surfaces. Hydration forces were not observed for either alumina substrates or alumina nanoparticles in both 10(-4) M KCl solution at pH 6.5 and 10(-2) M KCl at pH 10.2. Instead, strong attractive forces between the silicon nitride tip and the alumina (nanoparticles and substrate) were observed.

  15. The formation of liquid bridge in different operating modes of AFM

    Science.gov (United States)

    Wei, Zheng; Sun, Yan; Ding, WenXuan; Wang, ZaiRan

    2016-09-01

    The liquid bridge is one of the principal factors that cause artifacts in ambient-pressure atomic force microscope (AFM) images. Additionally, it is the main component of the adhesion force in ambient conditions. To understand the AFM imaging mechanism and the sample characteristics, it is essential to study the liquid bridge. This study interprets the physical mechanism involved in liquid bridge formation, which is composed of three different physical processes: the squeezing process, capillary condensation, and liquid film flow. We discuss the contributions of these three mechanisms to the volume and the capillary force of the liquid bridge in different AFM operation modes.

  16. Sensitivity of surface roughness parameters to changes in the density of scanning points in multi-scale AFM studies. Application to a biomaterial surface

    International Nuclear Information System (INIS)

    Mendez-Vilas, A.; Bruque, J.M.; Gonzalez-Martin, M.L.

    2007-01-01

    In the field of biomaterials surfaces, the ability of the atomic force microscope (AFM) to access the surface structure at unprecedented spatial (vertical and lateral) resolution, is helping in a better understanding on how topography affects the overall interaction of biological cells with the material surface. Since cells in a wide range of sizes are in contact with the biomaterial surface, a quantification of the surface structure in such a wide range of dimensional scales is needed. With the advent of the AFM, this can be routinely done in the lab. In this work, we show that even when it is clear that such a scale-dependent study is needed, AFM maps of the biomaterial surface taken at different scanning lengths are not completely consistent when they are taken at the same scanning resolution, as it is usually done: AFM images of different scanning areas have different point-to-point physical distances. We show that this effect influences the quantification of the average (R a ) and rms (R q ) roughness parameters determined at different length scales. This is the first time this inconsistency is reported and should be taken into account when roughness is measured in this way. Since differences will be in general in the range of nanometres, this is especially interesting for those processes involving the interaction of the biomaterial surface with small biocolloids as bacteria, while this effect should not represent any problems for larger animal cells

  17. Neutron detection using CR-39 and Atomic Force Microscopy (AFM)

    International Nuclear Information System (INIS)

    Vazquez L, C.; Fragoso, R.; Felix, R.; Golzarri, J.I.; Espinosa, G.; Castillo, F.

    2007-01-01

    AFM has been applied in many CR-39 track formation analyses. In this paper, the use of AFM in the neutron detection and analysis of the track formation is reported. The irradiation was made with an 1.5 GBq (0.5 Ci) 241 Am-Be neutron source, with and without a polyethylene radiator. The surface analysis was made to the CR-39 fresh material without irradiation, after the irradiation, and after a very short etching time. The results show important differences between the irradiation, with and without polyethylene radiator, and the latent tracks of the neutron in the CR-39 polycarbonate. The development of track formation after very short etching time and pits characterization were measured too using the AFM facilities. (Author)

  18. Noncontact AFM Imaging of Atomic Defects on the Rutile TiO2 (110) Surface

    DEFF Research Database (Denmark)

    Lauritsen, Jeppe Vang

    2015-01-01

    The atomic force microscope (AFM) operated in the noncontact mode (nc-AFM) offers a unique tool for real space, atomic-scale characterisation of point defects and molecules on surfaces, irrespective of the substrate being electrically conducting or non-conducting. The nc-AFM has therefore in rece...

  19. AFM topographies of densely packed nanoparticles: a quick way to determine the lateral size distribution by autocorrelation function analysis

    International Nuclear Information System (INIS)

    Fekete, L.; Kůsová, K.; Petrák, V.; Kratochvílová, I.

    2012-01-01

    The distribution of sizes is one of the basic characteristics of nanoparticles. Here, we propose a novel way to determine the lateral distribution of sizes from AFM topographies. Our algorithm is based on the autocorrelation function and can be applied both on topographies containing spatially separated and densely packed nanoparticles as well as on topographies of polycrystalline films. As no manual treatment is required, this algorithm can be easily automatable for batch processing. The algorithm works in principle with any kind of spatially mapped information (AFM current maps, optical microscope images, etc.), and as such has no size limitations. However, in the case of AFM topographies, the tip/sample convolution effects will be the factor limiting the smallest size to which the algorithm is applicable. Here, we demonstrate the usefulness of this algorithm on objects with sizes ranging between 20 nm and 1.5 μm.

  20. Fabrication of cone-shaped boron doped diamond and gold nanoelectrodes for AFM-SECM

    Energy Technology Data Exchange (ETDEWEB)

    Avdic, A; Lugstein, A; Bertagnolli, E [Solid State Electronics Institute, Vienna University of Technology, Floragasse 7, 1040 Vienna (Austria); Wu, M; Gollas, B [Competence Centre for Electrochemical Surface Technology, Viktor Kaplan Strasse 2, 2700 Wiener Neustadt (Austria); Pobelov, I; Wandlowski, T [Department of Chemistry and Biochemistry, University of Bern, Freiestrasse 3, 3012 Bern (Switzerland); Leonhardt, K; Denuault, G, E-mail: alois.lugstein@tuwien.ac.at [School of Chemistry, University of Southampton, Highfield, Southampton SO17 1BJ (United Kingdom)

    2011-04-08

    We demonstrate a reliable microfabrication process for a combined atomic force microscopy (AFM) and scanning electrochemical microscopy (SECM) measurement tool. Integrated cone-shaped sensors with boron doped diamond (BDD) or gold (Au) electrodes were fabricated from commercially available AFM probes. The sensor formation process is based on mature semiconductor processing techniques, including focused ion beam (FIB) machining, and highly selective reactive ion etching (RIE). The fabrication approach preserves the geometry of the original AFM tips resulting in well reproducible nanoscaled sensors. The feasibility and functionality of the fully featured tips are demonstrated by cyclic voltammetry, showing good agreement between the measured and calculated currents of the cone-shaped AFM-SECM electrodes.

  1. AFM study of structure influence on butterfly wings coloration

    OpenAIRE

    Dallaeva, Dinara; Tománek, Pavel

    2012-01-01

    This study describes the structural coloration of the butterfly Vanessa Atalanta wings and shows how the atomic force microscopy (AFM) can be applied to the study of wings morphology and wings surface behavior under the temperature. The role of the wings morphology in colors was investigated. Different colors of wings have different topology and can be identified by them. AFM in semi-contact mode was used to study the wings surface. The wing surface area, which is close to the butterfly body,...

  2. Transient Tip-Sample Interactions in High-Speed AFM Imaging of 3D nano structures

    NARCIS (Netherlands)

    Keyvani Janbahan, A.; Sadeghian Marnani, H; Goosen, H.; Keulen, F. van

    2015-01-01

    The maximum amount of repulsive force applied to the surface plays a very important role in damage of tip or sample in Atomic Force Microscopy(AFM). So far, many investigations have focused on peak repulsive forces in tapping mode AFM in steady state conditions. However, it is known that AFM could

  3. Implementing atomic force microscopy (AFM) for studying kinetics of gold nanoparticle's growth

    DEFF Research Database (Denmark)

    Georgiev, P.; Bojinova, A.; Kostova, B.

    2013-01-01

    In a novel experimental approach Atomic Force Microscopy (AFM) was applied as a tool for studying the kinetics of gold nanoparticle growth. The gold nanoparticles were obtained by classical Turkevich citrate synthesis at two different temperatures. From the analysis of AFM images during...... the synthesis process the nanoparticle s' sizes were obtained. To demonstrate the applicability and the reliability of the proposed experimental approach we studied the nanoparticles growth at two different temperatures by spectrophotometric measurements and compared them with the results from AFM experimental...

  4. AFM imaging of functionalized carbon nanotubes on biological membranes

    International Nuclear Information System (INIS)

    Lamprecht, C; Danzberger, J; Rangl, M; Gruber, H J; Hinterdorfer, P; Kienberger, F; Ebner, A; Liashkovich, I; Neves, V; Heister, E; Coley, H M; McFadden, J; Flahaut, E

    2009-01-01

    Multifunctional carbon nanotubes are promising for biomedical applications as their nano-size, together with their physical stability, gives access into the cell and various cellular compartments including the nucleus. However, the direct and label-free detection of carbon nanotube uptake into cells is a challenging task. The atomic force microscope (AFM) is capable of resolving details of cellular surfaces at the nanometer scale and thus allows following of the docking of carbon nanotubes to biological membranes. Here we present topographical AFM images of non-covalently functionalized single walled (SWNT) and double walled carbon nanotubes (DWNT) immobilized on different biological membranes, such as plasma membranes and nuclear envelopes, as well as on a monolayer of avidin molecules. We were able to visualize DWNT on the nuclear membrane while at the same time resolving individual nuclear pore complexes. Furthermore, we succeeded in localizing individual SWNT at the border of incubated cells and in identifying bundles of DWNT on cell surfaces by AFM imaging.

  5. Near-Field Spectroscopy with Nanoparticles Deposited by AFM

    Science.gov (United States)

    Anderson, Mark S.

    2008-01-01

    An alternative approach to apertureless near-field optical spectroscopy involving an atomic-force microscope (AFM) entails less complexity of equipment than does a prior approach. The alternative approach has been demonstrated to be applicable to apertureless near-field optical spectroscopy of the type using an AFM and surface enhanced Raman scattering (SERS), and is expected to be equally applicable in cases in which infrared or fluorescence spectroscopy is used. Apertureless near-field optical spectroscopy is a means of performing spatially resolved analyses of chemical compositions of surface regions of nanostructured materials. In apertureless near-field spectroscopy, it is common practice to utilize nanostructured probe tips or nanoparticles (usually of gold) having shapes and dimensions chosen to exploit plasmon resonances so as to increase spectroscopic-signal strengths. To implement the particular prior approach to which the present approach is an alternative, it is necessary to integrate a Raman spectrometer with an AFM and to utilize a special SERS-active probe tip. The resulting instrumentation system is complex, and the tasks of designing and constructing the system and using the system to acquire spectro-chemical information from nanometer-scale regions on a surface are correspondingly demanding.

  6. The LER/LWR metrology challenge for advance process control through 3D-AFM and CD-SEM

    Science.gov (United States)

    Faurie, P.; Foucher, J.; Foucher, A.-L.

    2009-12-01

    The continuous shrinkage in dimensions of microelectronic devices has reached such level, with typical gate length in advance R&D of less than 20nm combine with the introduction of new architecture (FinFET, Double gate...) and new materials (porous interconnect material, 193 immersion resist, metal gate material, high k materials...), that new process parameters have to be well understood and well monitored to guarantee sufficient production yield in a near future. Among these parameters, there are the critical dimensions (CD) associated to the sidewall angle (SWA) values, the line edge roughness (LER) and the line width roughness (LWR). Thus, a new metrology challenge has appeared recently and consists in measuring "accurately" the fabricated patterns on wafers in addition to measure the patterns on a repeatable way. Therefore, a great effort has to be done on existing techniques like CD-SEM, Scatterometry and 3D-AFM in order to develop them following the two previous criteria: Repeatability and Accuracy. In this paper, we will compare the 3D-AFM and CD-SEM techniques as a mean to measure LER and LWR on silicon and 193 resist and point out CD-SEM impact on the material during measurement. Indeed, depending on the material type, the interaction between the electron beam and the material or between the AFM tip and the material can vary a lot and subsequently can generate measurements bias. The first results tend to show that depending on CD-SEM conditions (magnification, number of acquisition frames) the final outputs can vary on a large range and therefore show that accuracy in such measurements are really not obvious to obtain. On the basis of results obtained on various materials that present standard sidewall roughness, we will show the limit of each technique and will propose different ways to improve them in order to fulfil advance roadmap requirements for the development of the next IC generation.

  7. Beyond topography - enhanced imaging of cometary dust with the MIDAS AFM

    Science.gov (United States)

    Bentley, M. S.; Torkar, K.; Jeszenszky, H.; Romstedt, J.

    2013-09-01

    The MIDAS atomic force microscope (AFM) onboard the Rosetta spacecraft is primarily designed to return the 3D shape and structure of cometary dust particles collected at comet 67P/Churyumov-Gerasimenko [1]. Commercial AFMs have, however, been further developed to measure many other sample properties. The possibilities to make such measurements with MIDAS are explored here.

  8. Visualising the Micro World of Chemical/Geochemical Interactions Using Atomic Force Microscopy (AFM)

    Energy Technology Data Exchange (ETDEWEB)

    Graham, G M; Sorbie, K S

    1997-12-31

    Scanning force microscopy, in particular AFM (Atomic Force Microscopy), provides a particular useful and interesting tool for the examination of surface structure at the near-atomic level. AFM is particularly well suited to the study of interactions at the surface in aqueous solutions using real time in-situ measurements. In this paper there is presented AFM images showing in situ crystal growth from supersaturated BaSO{sub 4} solutions onto the surface of barite. Growth structures in the form of spiral crystal growth features, presumably originating from screw dislocations, are illustrated. AFM images of novel scale crystal growth inhibition experiments are presented. Examination of the manner in which generically different species adsorb onto growth structures may help to explain mechanistic differences in the way which different inhibitor species perform against barium sulphate scale formation. Adsorption of polyacrylamide species onto mica surfaces have been viewed. The general utility of AFM to a number of other common surface interactions in oil field chemistry will be discussed. 17 refs., 3 figs.

  9. AFM-based identification of the dynamic properties of globular proteins: simulation study

    International Nuclear Information System (INIS)

    Kim, Deok Ho; Park, Jung Yul; Kim, Moon K.; Hong, Keum Shik

    2008-01-01

    Nowadays a mathematical model-based computational approach is getting more attention as an effective tool for understanding the mechanical behaviors of biological systems. To find the mechanical properties of the proteins required to build such a model, this paper investigates a real-time identification method based on an AFM nanomanipulation system. First, an AFM-based bio-characterization system is introduced. Second, a second-order time-varying linear model representing the interaction between an AFM cantilever and globular proteins in a solvent is presented. Finally, we address a real-time estimation method in which the results of AFM experiments are designed to be inputs of the state estimator proposed here. Our attention is restricted to a theoretical feasibility analysis of the proposed methodology. We simply set the mechanical properties of the particular protein such as mass, stiffness, and damping coefficient in the system model prior to running the simulation. Simulation results show very good agreement with the preset properties. We anticipate that the realization of the AFM-based bio-characterization system will also provide an experimental validation of the proposed identification procedure in the future. This methodology can be used to determine a model of protein motion for the purpose of computer simulation and for a real-time modification of protein deformation

  10. AFM-CMM integrated instrument user manual

    DEFF Research Database (Denmark)

    Marinello, Francesco; Bariani, Paolo

    This manual gives general important guidelines for a proper use of the integrated AFM-CMM instrument. More information can be collected reading: • N. Kofod Ph.D thesis [1]; • P. Bariani Ph.D thesis [2]; • Dualscope DME 95-200 operation manuals [3]; • SPIP help [4] • Stitching software user manual...

  11. Sensing inhomogeneous mechanical properties of human corneal Descemet's membrane with AFM nano-indentation.

    Science.gov (United States)

    Di Mundo, Rosa; Recchia, Giuseppina; Parekh, Mohit; Ruzza, Alessandro; Ferrari, Stefano; Carbone, Giuseppe

    2017-10-01

    The paper describes a highly space-resolved characterization of the surface mechanical properties of the posterior human corneal layer (Descemet's membrane). This has been accomplished with Atomic Force Microscopy (AFM) nano-indentation by using a probe with a sharp tip geometry. Results indicate that the contact with this biological tissue in liquid occurs with no (or very low) adhesion. More importantly, under the same operating conditions, a broad distribution of penetration depth can be measured on different x-y positions of the tissue surface, indicating a high inhomogeneity of surface stiffness, not yet clearly reported in the literature. An important contribution to such inhomogeneity should be ascribed to the discontinuous nature of the collagen/proteoglycans fibers matrix tissue, as can be imaged by AFM when the tissue is semi-dry. Using classical contact mechanics calculations adapted to the specific geometry of the tetrahedral tip it has been found that the elastic modulus E of the material in the very proximity of the surface ranges from 0.23 to 2.6 kPa. Copyright © 2017 Elsevier Ltd. All rights reserved.

  12. Topographical heterogeneity in transparent PVA hydrogels studied by AFM

    Energy Technology Data Exchange (ETDEWEB)

    Pramanick, Ashit Kumar; Gupta, Siddhi, E-mail: siddhigupta@nmlindia.org; Mishra, Trilochan; Sinha, Arvind

    2012-02-01

    Physically crosslinked poly (vinyl alcohol) (PVA) hydrogels have a wide range of biomedical applications. Transparent and stable PVA hydrogels synthesized by freeze-thawing method are potential candidates to be used as tissue engineering scaffolds provided they exhibit suitable topographical roughness and surface energy. The effect of processing parameters i.e., polymer concentration and number of freeze-thaw cycles on the resulting topography of the freeze-thawed transparent hydrogels has been studied and quantified using non-contact mode of an atomic force microscope (AFM) and image analysis. Simultaneously captured phase contrast images have revealed significant information about morphological changes in the topographical features and crystallinity of the hydrogels. Topographical roughness was found to decrease as a function of number of freeze-thaw cycles.

  13. An AFM-based pit-measuring method for indirect measurements of cell-surface membrane vesicles

    International Nuclear Information System (INIS)

    Zhang, Xiaojun; Chen, Yuan; Chen, Yong

    2014-01-01

    Highlights: • Air drying induced the transformation of cell-surface membrane vesicles into pits. • An AFM-based pit-measuring method was developed to measure cell-surface vesicles. • Our method detected at least two populations of cell-surface membrane vesicles. - Abstract: Circulating membrane vesicles, which are shed from many cell types, have multiple functions and have been correlated with many diseases. Although circulating membrane vesicles have been extensively characterized, the status of cell-surface membrane vesicles prior to their release is less understood due to the lack of effective measurement methods. Recently, as a powerful, micro- or nano-scale imaging tool, atomic force microscopy (AFM) has been applied in measuring circulating membrane vesicles. However, it seems very difficult for AFM to directly image/identify and measure cell-bound membrane vesicles due to the similarity of surface morphology between membrane vesicles and cell surfaces. Therefore, until now no AFM studies on cell-surface membrane vesicles have been reported. In this study, we found that air drying can induce the transformation of most cell-surface membrane vesicles into pits that are more readily detectable by AFM. Based on this, we developed an AFM-based pit-measuring method and, for the first time, used AFM to indirectly measure cell-surface membrane vesicles on cultured endothelial cells. Using this approach, we observed and quantitatively measured at least two populations of cell-surface membrane vesicles, a nanoscale population (<500 nm in diameter peaking at ∼250 nm) and a microscale population (from 500 nm to ∼2 μm peaking at ∼0.8 μm), whereas confocal microscopy only detected the microscale population. The AFM-based pit-measuring method is potentially useful for studying cell-surface membrane vesicles and for investigating the mechanisms of membrane vesicle formation/release

  14. AFM plough YBCO micro bridges: substrate effects

    CSIR Research Space (South Africa)

    Elkaseh, A

    2009-04-01

    Full Text Available AFM nanolithography was used as a novel cutting technique to define micro-size YBCO superconducting constrictions. Researchers studied the substrate effects on MgO and STO substrates and showed that the observed Shapiro steps from the bridges on STO...

  15. Application of Contact Mode AFM to Manufacturing Processes

    Science.gov (United States)

    Giordano, Michael A.; Schmid, Steven R.

    A review of the application of contact mode atomic force microscopy (AFM) to manufacturing processes is presented. A brief introduction to common experimental techniques including hardness, scratch, and wear testing is presented, with a discussion of challenges in the extension of manufacturing scale investigations to the AFM. Differences between the macro- and nanoscales tests are discussed, including indentation size effects and their importance in the simulation of processes such as grinding. The basics of lubrication theory are presented and friction force microscopy is introduced as a method of investigating metal forming lubrication on the nano- and microscales that directly simulates tooling/workpiece asperity interactions. These concepts are followed by a discussion of their application to macroscale industrial manufacturing processes and direct correlations are made.

  16. Spin dynamics and quantum tunneling in Fe8 nanomagnet and in AFM rings by NMR

    Science.gov (United States)

    Baek, Seung-Ho

    In this thesis, our main interest has been to investigate the spin dynamics and quantum tunneling in single molecule magnets (SMMs). For this we have selected two different classes of SMMs: a ferromagnetic total high spin S = 10 cluster Fe8 and antiferromagnetic (AFM) ring-type clusters. For Fe8, our efforts have been devoted to the investigation of the quantum tunneling of magnetization in the very low temperature region. The most remarkable experimental finding in Fe8 is that the nuclear spin-lattice relaxation rate (1/T1) at low temperatures takes place via strong collision mechanism, and thus it allows to measure directly the tunneling rate vs. T and H for the first time. For AFM rings, we have shown that 1/T1 probes the thermal fluctuations of the magnetization in the intermediate temperature range. We find that the fluctuations are dominated by a single characteristic frequency which has a power-law T-dependence indicative of fluctuations due to electron-acoustic phonon interactions.

  17. Spin Dynamics and Quantum Tunneling in Fe8 Nanomagnet and in AFM Rings by NMR

    International Nuclear Information System (INIS)

    Seung-Ho-Baek

    2004-01-01

    In this thesis, our main interest has been to investigate the spin dynamics and quantum tunneling in single molecule magnets (SMMs), For this we have selected two different classes of SMMs: a ferrimagnetic total high spin S = 10 cluster Fe8 and antiferromagnetic (AFM) ring-type clusters. For Fe8, our efforts have been devoted to the investigation of the quantum tunneling of magnetization in the very low temperature region. The most remarkable experimental finding in Fe8 is that the nuclear spin-lattice relaxation rate (1/T l ) at low temperatures takes place via strong collision mechanism, and thus it allows to measure directly the tunneling rate vs T and H for the first time. For AFM rings, we have shown that 1/T l probes the thermal fluctuations of the magnetization in the intermediate temperature range. We find that the fluctuations are dominated by a single characteristic frequency which has a power law T-dependence indicative of fluctuations due to electron-acoustic phonon interactions

  18. AFM Study of Structure Influence on Butterfly Wings Coloration

    Directory of Open Access Journals (Sweden)

    Dinara Sultanovna Dallaeva

    2012-01-01

    Full Text Available This study describes the structural coloration of the butterfly Vanessa Atalanta wings and shows how the atomic force microscopy (AFM can be applied to the study of wings morphology and wings surface behavior under the temperature. The role of the wings morphology in colors was investigated. Different colors of wings have different topology and can be identified by them. AFM in semi-contact mode was used to study the wings surface. The wing surface area, which is close to the butterfly body, has shiny brown color and the peak of surface roughness is about 600 nm. The changing of morphology at different temperatures is shown.

  19. Systematic UHV-AFM experiments on Na nano-particles and nano-structures in NaCl

    OpenAIRE

    Sugonyako, A.V.; Turkin, A.A.; Gaynutdinov, R.; Vainshtein, D.I.; Hartog, H.W. den; Bukharaev, A.A.

    2005-01-01

    Results of systematic AFM (atomic force microscopy) experiments on heavily and moderatly irradiated NaCl samples are presented. The sodium nanoparticles and structures of nanoparticles are poduced in sodium chloride during irradiation. The AFM images of the nanoparticles have been obtained in ultra high vacuum (UHV) in the non-contact mode with an Omicron UHV AFM/STM system. The sizes and arrangements of the observed particles depend on the irradiation conditions. The melting behaviour of the...

  20. Charge storage in mesoscopic graphitic islands fabricated using AFM bias lithography

    Energy Technology Data Exchange (ETDEWEB)

    Kurra, Narendra; Basavaraja, S; Kulkarni, G U [Chemistry and Physics of Materials Unit and DST Unit on Nanoscience, Jawaharlal Nehru Centre for Advanced Scientific Research, Jakkur PO, Bangalore 560 064 (India); Prakash, Gyan; Fisher, Timothy S; Reifenberger, Ronald G, E-mail: kulkarni@jncasr.ac.in, E-mail: reifenbr@purdue.edu [Birck Nanotechnology Center, Purdue University, West Lafayette, IN 47907 (United States)

    2011-06-17

    Electrochemical oxidation and etching of highly oriented pyrolytic graphite (HOPG) has been achieved using biased atomic force microscopy (AFM) lithography, allowing patterns of varying complexity to be written into the top layers of HOPG. The graphitic oxidation process and the trench geometry after writing were monitored using intermittent contact mode AFM. Electrostatic force microscopy reveals that the isolated mesoscopic islands formed during the AFM lithography process become positively charged, suggesting that they are laterally isolated from the surrounding HOPG substrate. The electrical transport studies of these laterally isolated finite-layer graphitic islands enable detailed characterization of electrical conduction along the c-direction and reveal an unexpected stability of the charged state. Utilizing conducting-atomic force microscopy, the measured I(V) characteristics revealed significant non-linearities. Micro-Raman studies confirm the presence of oxy functional groups formed during the lithography process.

  1. Finite element modeling of trolling-mode AFM.

    Science.gov (United States)

    Sajjadi, Mohammadreza; Pishkenari, Hossein Nejat; Vossoughi, Gholamreza

    2018-06-01

    Trolling mode atomic force microscopy (TR-AFM) has overcome many imaging problems in liquid environments by considerably reducing the liquid-resonator interaction forces. The finite element model of the TR-AFM resonator considering the effects of fluid and nanoneedle flexibility is presented in this research, for the first time. The model is verified by ABAQUS software. The effect of installation angle of the microbeam relative to the horizon and the effect of fluid on the system behavior are investigated. Using the finite element model, frequency response curve of the system is obtained and validated around the frequency of the operating mode by the available experimental results, in air and liquid. The changes in the natural frequencies in the presence of liquid are studied. The effects of tip-sample interaction on the excitation of higher order modes of the system are also investigated in air and liquid environments. Copyright © 2018 Elsevier B.V. All rights reserved.

  2. Electrospray deposition from fountain pen AFM probes

    NARCIS (Netherlands)

    Geerlings, J.; Sarajlic, Edin; Berenschot, Johan W.; Abelmann, Leon; Tas, Niels Roelof

    2012-01-01

    In this paper we present for the first time electrospraying from fountain pen probes. By using electrospray contactless deposition in an AFM setup becomes possible. Experiments on a dedicated setup were carried out as first step towards this goal. Spraying from 8 and 2 µm apertures was observed. For

  3. Fast and controlled fabrication of porous graphene oxide: application of AFM tapping for mechano-chemistry

    Science.gov (United States)

    Chu, Liangyong; Korobko, Alexander V.; Bus, Marcel; Boshuizen, Bart; Sudhölter, Ernst J. R.; Besseling, Nicolaas A. M.

    2018-05-01

    This paper describes a novel method to fabricate porous graphene oxide (PGO) from GO by exposure to oxygen plasma. Compared to other methods to fabricate PGO described so far, e.g. the thermal and steam etching methods, oxygen plasma etching method is much faster. We studied the development of the porosity with exposure time using atomic force microscopy (AFM). It was found that the development of PGO upon oxygen-plasma exposure can be controlled by tapping mode AFM scanning using a Si tip. AFM tapping stalls the growth of pores upon further plasma exposure at a level that coincides with the fraction of sp2 carbons in the GO starting material. We suggest that AFM tapping procedure changes the bond structure of the intermediate PGO structure, and these stabilized PGO structures cannot be further etched by oxygen plasma. This constitutes the first report of tapping AFM as a tool for local mechano-chemistry.

  4. AFM of metallic nano-particles and nano-structures in heavily irradiated NaCl

    NARCIS (Netherlands)

    Gaynutdinov, R; Vainshtein, DI; Hak, SJ; Tolstikhina, A; Den Hartog, HW

    2003-01-01

    AFM investigations are reported for heavily, electron irradiated NaCl crystals in ultra high vacuum (UHV) in the non-contact mode-with an UHV AFM/STM Omicron system. To avoid chemical reactions between the radiolytic Na and oxygen and water, the irradiated samples were cleaved and prepared for the

  5. Spontaneous aggregation of humic acid observed with AFM at different pH.

    Science.gov (United States)

    Colombo, Claudio; Palumbo, Giuseppe; Angelico, Ruggero; Cho, Hyen Goo; Francioso, Ornella; Ertani, Andrea; Nardi, Serenella

    2015-11-01

    Atomic force microscopy in contact (AFM-C) mode was used to investigate the molecular dynamics of leonardite humic acid (HA) aggregate formed at different pH values. HA nanoparticles dispersed at pH values ranging from 2 to 12 were observed on a mica surface under dry conditions. The most clearly resolved and well-resulted AFM images of single particle were obtained at pH 5, where HA appeared as supramolecular particles with a conic shape and a hole in the centre. Those observations suggested that HA formed under these conditions exhibited a pseudo-amphiphilic nature, with secluded hydrophobic domains and polar subunits in direct contact with hydrophilic mica surface. Based on molecular simulation methods, a lignin-carbohydrate complex (LCC) model was proposed to explain the HA ring-like morphology. The LCC model optimized the parameters of β-O-4 linkages between 14 units of 1-4 phenyl propanoid, and resulted in an optimized structure comprising 45-50 linear helical molecules looped spirally around a central cavity. Those results added new insights on the adsorption mechanism of HA on polar surfaces as a function of pH, which was relevant from the point of view of natural aggregation in soil environment. Copyright © 2015 Elsevier Ltd. All rights reserved.

  6. Characteristics of fracture during the approach process and wear mechanism of a silicon AFM tip

    International Nuclear Information System (INIS)

    Chung, Koo-Hyun; Lee, Yong-Ha; Kim, Dae-Eun

    2005-01-01

    The wear of an atomic force microscope (AFM) tip is one of the crucial issues in AFM as well as in other probe-based applications. In this work, wear tests under extremely low normal load using an AFM were conducted. Also, in order to understand the nature of silicon tip wear, the wear characteristics of crystal silicon and amorphous silicon oxide layer were investigated by a high-resolution transmission electron microscope (HRTEM). It was found that fracture of the tip readily occurred due to impact during the approach process. Experimental results showed that the impact should be below 0.1 nN s to avoid significant fracture of the tip. Also, it was observed that wear of the amorphous layer, formed at the end of the tip, occurred at the initial stage of the silicon tip damage process. Based on Archard's wear law, the wear coefficient of the amorphous layer was in the range of 0.009-0.014. As for the wear characteristics of the silicon tip, it was shown that wear occurred gradually under light normal load and the wear rate decreased with increase in the sliding distance. As for the wear mechanism of the silicon tip, oxidation wear was identified to be the most significant. It was shown that the degree of oxidation was higher under high normal load and in a nitrogen environment, oxidation of the silicon tip was reduced

  7. Superparamagnetism in AFM Cr2O3 nanoparticles

    International Nuclear Information System (INIS)

    Tobia, D.; Winkler, E.L.; Zysler, R.D.; Granada, M.; Troiani, H.E.

    2010-01-01

    In this work we report the size effects on the magnetic properties of AFM Cr 2 O 3 nanoparticles. From transmission electron microscopy we determined that the system presents high crystallinity and narrow lognormal size distribution centred at = 7.8 nm with σ = 0.3. The magnetic properties of the nanoparticles were studied by magnetization and electron paramagnetic resonance (EPR) experiments. By EPR spectroscopy we established that the AFM order temperature, T N , shifted to ∼270 K when the size is reduced (T N (Bulk) ∼ 308 K). From the zero-field-cooling and the field-cooling magnetization curves we determined the blocking temperature T B = 28 K. Below T B the system presents exchange bias effect. We discuss the results by using recent models in terms of the internal magnetic structures of the nanoparticles.

  8. Comparison of different aminofunctionalization strategies for attachment of single antibodies to AFM cantilevers

    Energy Technology Data Exchange (ETDEWEB)

    Ebner, Andreas [Institute of Biophysics, University of Linz, 4040 Linz (Austria)], E-mail: andreas.ebner@jku.at; Hinterdorfer, Peter; Gruber, Hermann J. [Institute of Biophysics, University of Linz, 4040 Linz (Austria)

    2007-10-15

    Atomic force microscopy (AFM) has developed into a key technique for elucidation of biological systems on the single molecular level. In particular, molecular recognition force microscopy has proven to be a powerful tool for the investigation of biological interactions under near physiological conditions. For this purpose, ligands are tethered to AFM tips and the interaction forces with cognate receptors on the sample surface are measured with pico-Newton accuracy. In the first step of tip functionalization, amino groups are typically introduced on the initially inert AFM tip. Several methods have been developed to reproducibly adjust the desired low density of amino groups on the tip surface, i.e. esterification with ethanolamine, gas-phase silanization with aminopropyl-triethoxysilane (APTES), or treatment with aminophenyl-trimethoxysilane (APhS) in toluene solution. In the present study, the usefulness of these methods for attachments of antibodies to AFM tips was characterized by a standardized test system, in which biotinylated IgG was bound to the tip and a dense monolayer of avidin on mica served as test sample. All three methods of aminofunctionalization were found fully satisfactory for attachment of single antibodies to AFM tips, only in a parallel macroscopic assay on silicon nitride chips a minor difference was found in that APTES appeared to yield a slightly lower surface density of amino groups.

  9. Immobilizing live Escherichia coli for AFM studies of surface dynamics

    International Nuclear Information System (INIS)

    Lonergan, N.E.; Britt, L.D.; Sullivan, C.J.

    2014-01-01

    Atomic force microscopy (AFM) is a probe-based technique that permits high resolution imaging of live bacterial cells. However, stably immobilizing cells to withstand the probe-based lateral forces remains an obstacle in AFM mediated studies, especially those of live, rod shaped bacteria in nutrient media. Consequently, AFM has been under-utilized in the research of bacterial surface dynamics. The aim of the current study was to immobilize a less adherent Escherichia coli strain in a method that both facilitates AFM imaging in nutrient broth and preserves overall cell viability. Immobilization reagents and buffers were systematically evaluated and the cell membrane integrity was monitored in all sample preparations. As expected, the biocompatible gelatin coated surfaces facilitated stable cell attachment in lower ionic strength buffers, yet poorly immobilized cells in higher ionic strength buffers. In comparison, poly-L-lysine surfaces bound cells in both low and high ionic strength buffers. The benefit of the poly-L-lysine binding capacity was offset by the compromised membrane integrity exhibited by cells on poly-L-lysine surfaces. However, the addition of divalent cations and glucose to the immobilization buffer was found to mitigate this unfavorable effect. Ultimately, immobilization of E. coli cells on poly-L-lysine surfaces in a lower ionic strength buffer supplemented with Mg 2+ and Ca 2+ was determined to provide optimal cell attachment without compromising the overall cell viability. Cells immobilized in this method were stably imaged in media through multiple division cycles. Furthermore, permeability assays indicated that E. coli cells recover from the hypoosmotic stress caused by immobilization in low ionic strength buffers. Taken together, this data suggests that stable immobilization of viable cells on poly-L-lysine surfaces can be accomplished in lower ionic strength buffers that are supplemented with divalent cations for membrane stabilization while

  10. AFM study of growth of Bi2Sr2-xLaxCuO6 thin films

    International Nuclear Information System (INIS)

    Haitao Yang; Hongjie Tao; Yingzi Zhang; Duogui Yang; Lin Li; Zhongxian Zhao

    1997-01-01

    c-axis-oriented Bi 2 Sr 1.6 La 0.4 CuO 6 thin films deposited on flat planes of (100)SrTiO 3 , (100)LaAlO 3 and (100)MgO substrates and vicinal planes (off-angle ∼ 6 deg.) of SrTiO 3 substrates by RF magnetron sputtering were studied by atomic force microscopy (AFM). T c of these films reached 29 K. Film thickness ranged from 15 nm to 600 nm. Two typical growth modes have been observed. AFM images of thin films on flat planes of substrates showed a terraced-island growth mode. By contrast, Bi-2201 thin films on vicinal planes of substrates showed a step-flow growth mode. The growth unit is a half-unit-cell in the c-axis for both growth modes. No example of spiral growth, which was thought to be the typical structure of YBCO thin films, was found in either of these kinds of thin films. (author)

  11. AFM characterization of nonwoven material functionalized by ZnO sputter coating

    International Nuclear Information System (INIS)

    Deng Bingyao; Yan Xiong; Wei Qufu; Gao Weidong

    2007-01-01

    Sputter coatings provide new approaches to the surface functionalization of textile materials. In this study, polyethylene terephthalate (PET) nonwoven material was used as a substrate for creating functional nanostructures on the fiber surfaces. A magnetron sputter coating was used to deposit functional zinc oxide (ZnO) nanostructures onto the nonwoven substrate. The evolution of the surface morphology of the fibers in the nonwoven web was examined using atomic force microscopy (AFM). The AFM observations revealed a significant difference in the morphology of the fibers before and after the sputter coating. The AFM images also indicated the effect of the sputtering conditions on the surface morphology of the fibers. The increase in the sputtering time led to the growth of the ZnO grains on the fiber surfaces. The higher pressure in the sputtering chamber could cause the formation of larger grains on the fiber surfaces. The higher power used also generated larger grains on the fiber surfaces

  12. Systematic UHV-AFM experiments on Na nano-particles and nano-structures in NaCl

    NARCIS (Netherlands)

    Sugonyako, A.V.; Turkin, A.A.; Gaynutdinov, R.; Vainshtein, D.I.; Hartog, H.W. den; Bukharaev, A.A.

    2005-01-01

    Results of systematic AFM (atomic force microscopy) experiments on heavily and moderatly irradiated NaCl samples are presented. The sodium nanoparticles and structures of nanoparticles are poduced in sodium chloride during irradiation. The AFM images of the nanoparticles have been obtained in ultra

  13. Analysis of long- and short-range contribution to adhesion work in cardiac fibroblasts: An atomic force microscopy study

    Energy Technology Data Exchange (ETDEWEB)

    Sbaizero, O., E-mail: sbaizero@units.it [Department of Engineering and Architecture, University of Trieste (Italy); University of Colorado Cardiovascular Institute, University of Colorado Denver, Aurora (United States); DelFavero, G. [Department of Engineering and Architecture, University of Trieste (Italy); Martinelli, V. [International Center for Genetic Engineering and Biotechnology, Trieste (Italy); Long, C.S.; Mestroni, L. [University of Colorado Cardiovascular Institute, University of Colorado Denver, Aurora (United States)

    2015-04-01

    Atomic force microscopy (AFM) for single-cell force spectroscopy (SCFS) and Poisson statistic were used to analyze the detachment work recorded during the removal of gold-covered microspheres from cardiac fibroblasts. The effect of Cytochalasin D, a disruptor of the actin cytoskeleton, on cell adhesion was also tested. The adhesion work was assessed using a Poisson analysis also derived from single-cell force spectroscopy retracting curves. The use of Poisson analysis to get adhesion work from AFM curves is quite a novel method, and in this case, proved to be effective to study the short-range and long-range contributions to the adhesion work. This method avoids the difficult identification of minor peaks in the AFM retracting curves by creating what can be considered an average adhesion work. Even though the effect of actin depolymerisation is well documented, its use revealed that control cardiac fibroblasts (CT) exhibit a work of adhesion at least 5 times higher than that of the Cytochalasin treated cells. However, our results indicate that in both cells short-range and long-range contributions to the adhesion work are nearly equal and the same heterogeneity index describes both cells. Therefore, we infer that the different adhesion behaviors might be explained by the presence of fewer membrane adhesion molecules available at the AFM tip–cell interface under circumstances where the actin cytoskeleton has been disrupted. - Highlights: • AFM force–deformation curve was used to characterize the cardiac fibroblast adhesion behavior. • The amount and nature of adhesion were assessed using a Poisson analysis applied to the AFM curve. • The work of adhesion for control cells was about four times higher than that of the Cyt-D treated cells. • Short- and long-range contributions to adhesion are nearly equal for both control and treated cells.

  14. AFM cantilever with in situ renewable mercury microelectrode

    NARCIS (Netherlands)

    Schön, Peter Manfred; Geerlings, J.; Tas, Niels Roelof; Sarajlic, Edin

    2013-01-01

    We report here first results obtained on a novel, in situ renewable mercury microelectrode integrated into an atomic force microscopy (AFM) cantilever. Our approach is based on a fountain pen probe with appropriate dimensions enabling reversible filling with(nonwetting) mercury under changing the

  15. Nanopuller-open data acquisition platform for AFM force spectroscopy experiments

    International Nuclear Information System (INIS)

    Pawlak, Konrad; Strzelecki, Janusz

    2016-01-01

    Atomic Force Microscope (AFM) is a widely used tool in force spectroscopy studies. Presently, this instrument is accessible from numerous vendors, albeit commercial solutions are expensive and almost always hardware and software closed. Approaches for open setups were published, as with modern low cost and readily available piezoelectric actuators, data acquisition interfaces and optoelectronic components building such force spectroscopy AFM is relatively easy. However, suitable software to control such laboratory made instrument was not released. Developing it in the lab requires significant time and effort. Our Nanopuller software described in this paper is intended to eliminate this obstacle. With only minimum adjustments this program can be used to control and acquire data with any suitable National Instruments universal digital/analog interface and piezoelectric actuator analog controller, giving significant freedom and flexibility in designing force spectroscopy experiment. Since the full code, written in a graphical LabVIEW environment is available, our Nanopuller can be easily customized. In this paper we describe the program and test its performance in controlling different setups. Successful and accurate force curve acquisition for standard samples (single molecules of I27O reference titin polyprotein and DNA as well as red blood cells) is shown. - Highlights: • We created open data acquisition software for performing Atomic Force Microscopy force measurements with custom laboratory made setups. • The software allows large flexibility in atomic force microscope design with minimum adjustment necessary. • The software is written in LabVIEW, allowing easy customization. • We successfully tested the program on two different hardware configurations by stretching single macromolecules and indenting cells.

  16. Nanopuller-open data acquisition platform for AFM force spectroscopy experiments

    Energy Technology Data Exchange (ETDEWEB)

    Pawlak, Konrad; Strzelecki, Janusz

    2016-05-15

    Atomic Force Microscope (AFM) is a widely used tool in force spectroscopy studies. Presently, this instrument is accessible from numerous vendors, albeit commercial solutions are expensive and almost always hardware and software closed. Approaches for open setups were published, as with modern low cost and readily available piezoelectric actuators, data acquisition interfaces and optoelectronic components building such force spectroscopy AFM is relatively easy. However, suitable software to control such laboratory made instrument was not released. Developing it in the lab requires significant time and effort. Our Nanopuller software described in this paper is intended to eliminate this obstacle. With only minimum adjustments this program can be used to control and acquire data with any suitable National Instruments universal digital/analog interface and piezoelectric actuator analog controller, giving significant freedom and flexibility in designing force spectroscopy experiment. Since the full code, written in a graphical LabVIEW environment is available, our Nanopuller can be easily customized. In this paper we describe the program and test its performance in controlling different setups. Successful and accurate force curve acquisition for standard samples (single molecules of I27O reference titin polyprotein and DNA as well as red blood cells) is shown. - Highlights: • We created open data acquisition software for performing Atomic Force Microscopy force measurements with custom laboratory made setups. • The software allows large flexibility in atomic force microscope design with minimum adjustment necessary. • The software is written in LabVIEW, allowing easy customization. • We successfully tested the program on two different hardware configurations by stretching single macromolecules and indenting cells.

  17. In Situ STM and AFM of the Copper Protein Pseudomonas Aeruginosa Azurin

    DEFF Research Database (Denmark)

    Friis, Esben P.; Andersen, Jens Enevold Thaulov; Madsen, L.L.

    1997-01-01

    to gold and facile electron tunnel routes between this group and the copper atom. Azurin adsorbed on Au(111) can be imaged to molecular resolution by in situ STM and shows regular arrays of individual structures corresponding well to the known molecular size of azurin. The current falls off approximately...... exponentially with increasing distance with a decay constant of 0.4–0.5 Å−1. In comparison in situ AFM shows structures laterally convoluted with the tip while the vertical extension is in the same range as the structural size of azurin. The results are of interest in relation to electron tunnel mechanisms...

  18. Spin Dynamics and Quantum Tunneling in Fe8 Nanomagnet and in AFM Rings by NMR

    Energy Technology Data Exchange (ETDEWEB)

    Ho-Baek, Seung [Iowa State Univ., Ames, IA (United States)

    2004-01-01

    In this thesis, our main interest has been to investigate the spin dynamics and quantum tunneling in single molecule magnets (SMMs), For this we have selected two different classes of SMMs: a ferrimagnetic total high spin S = 10 cluster Fe8 and antiferromagnetic (AFM) ring-type clusters. For Fe8, our efforts have been devoted to the investigation of the quantum tunneling of magnetization in the very low temperature region. The most remarkable experimental finding in Fe8 is that the nuclear spin-lattice relaxation rate (1/T{sub l}) at low temperatures takes place via strong collision mechanism, and thus it allows to measure directly the tunneling rate vs T and H for the first time. For AFM rings, we have shown that 1/T{sub l} probes the thermal fluctuations of the magnetization in the intermediate temperature range. We find that the fluctuations are dominated by a single characteristic frequency which has a power law T-dependence indicative of fluctuations due to electron-acoustic phonon interactions.

  19. A fully-automated neural network analysis of AFM force-distance curves for cancer tissue diagnosis

    Science.gov (United States)

    Minelli, Eleonora; Ciasca, Gabriele; Sassun, Tanya Enny; Antonelli, Manila; Palmieri, Valentina; Papi, Massimiliano; Maulucci, Giuseppe; Santoro, Antonio; Giangaspero, Felice; Delfini, Roberto; Campi, Gaetano; De Spirito, Marco

    2017-10-01

    Atomic Force Microscopy (AFM) has the unique capability of probing the nanoscale mechanical properties of biological systems that affect and are affected by the occurrence of many pathologies, including cancer. This capability has triggered growing interest in the translational process of AFM from physics laboratories to clinical practice. A factor still hindering the current use of AFM in diagnostics is related to the complexity of AFM data analysis, which is time-consuming and needs highly specialized personnel with a strong physical and mathematical background. In this work, we demonstrate an operator-independent neural-network approach for the analysis of surgically removed brain cancer tissues. This approach allowed us to distinguish—in a fully automated fashion—cancer from healthy tissues with high accuracy, also highlighting the presence and the location of infiltrating tumor cells.

  20. FRAME (Force Review Automation Environment): MATLAB-based AFM data processor.

    Science.gov (United States)

    Partola, Kostyantyn R; Lykotrafitis, George

    2016-05-03

    Data processing of force-displacement curves generated by atomic force microscopes (AFMs) for elastic moduli and unbinding event measurements is very time consuming and susceptible to user error or bias. There is an evident need for consistent, dependable, and easy-to-use AFM data processing software. We have developed an open-source software application, the force review automation environment (or FRAME), that provides users with an intuitive graphical user interface, automating data processing, and tools for expediting manual processing. We did not observe a significant difference between manually processed and automatically processed results from the same data sets. Copyright © 2016 Elsevier Ltd. All rights reserved.

  1. Ultra-high aspect ratio replaceable AFM tips using deformation-suppressed focused ion beam milling

    DEFF Research Database (Denmark)

    Savenko, Alexey; Yildiz, Izzet; Petersen, Dirch Hjorth

    2013-01-01

    Fabrication of ultra-high aspect ratio exchangeable and customizable tips for atomic force microscopy (AFM) using lateral focused ion beam (FIB) milling is presented. While on-axis FIB milling does allow high aspect ratio (HAR) AFM tips to be defined, lateral milling gives far better flexibility...

  2. Adhesion experiments using an AFM-Parameters of influence

    NARCIS (Netherlands)

    Dos Santos Ferreira, O.; Gelinck, E.R.M.; Graaf, D. de; Fischer, H.

    2010-01-01

    Adhesion measurements were performed by AFM (Atomic Force Microscopy). It was shown that many parameters need to be controlled in order to provide reproducible and quantitative results. Adhesion forces were shown to depend on combination of materials characteristics and testing geometry as well as

  3. A facile alternative technique for large-area graphene transfer via sacrificial polymer

    Directory of Open Access Journals (Sweden)

    Eric Auchter

    2017-12-01

    Full Text Available A novel method of transferring large-area graphene sheets onto a variety of substrates using Formvar (polyvinyl formal is presented. Due to the ease at which formvar can be dissolved in chloroform this method allows for a consistent, a clean, and a more rapid transfer than other techniques including the PMMA assisted one. This novel transfer method is demonstrated by transferring large-area graphene onto a range of substrates including commercial TEM grids, silicon dioxide and glass. Raman spectroscopy was used to confirm the presence of graphene and characterize the morphological properties of the large-area sheets. SEM and AFM analyses demonstrated the effectiveness of our rapid transfer technique for clean crystalline large-area graphene sheets. The removal of the sacrificial polymer was found to be one to two orders of magnitude faster than PMMA methods. Ultimately this facile transfer technique offers new opportunities for a wide range of applications for large-area graphene through the utilization of a new sacrificial polymer.

  4. Optimization of Easy Atomic Force Microscope (ezAFM) Controls for Semiconductor Nanostructure Profiling

    Science.gov (United States)

    2017-09-01

    ARL-MR-0965 ● SEP 2017 US Army Research Laboratory Optimization of Easy Atomic Force Microscope (ezAFM) Controls for... Optimization of Easy Atomic Force Microscope (ezAFM) Controls for Semiconductor Nanostructure Profiling by Satwik Bisoi Science and...REPORT TYPE Memorandum Report 3. DATES COVERED (From - To) 2017 July 05–2017 August 18 4. TITLE AND SUBTITLE Optimization of Easy Atomic Force

  5. Self-assembled silver nanoparticles monolayers on mica-AFM, SEM, and electrokinetic characteristics

    International Nuclear Information System (INIS)

    Oćwieja, Magdalena; Morga, Maria; Adamczyk, Zbigniew

    2013-01-01

    A monodisperse silver particle suspension was produced by a chemical reduction method in an aqueous medium using sodium citrate. The average particle size determined by dynamic light scattering (DLS), transmission electron microscopy (TEM), and atomic force microscopy (AFM) was 28.5 nm. The DLS measurements confirmed that the suspension was stable for the ionic strength up to 3 × 10 −2 M NaCl. The electrophoretic mobility measurements revealed that the electrokinetic charge of particles was negative for pH range 3–10, assuming −50 e for pH = 9 and 0.01 M NaCl. Using the suspension, silver particle monolayers on mica modified by poly(allylamine hydrochloride) were produced under diffusion-controlled transport. Monolayer coverage, quantitatively determined by AFM and SEM, was regulated within broad limits by adjusting the nanoparticle deposition time. This allowed one to uniquely express the zeta potential of silver monolayers, determined by the in situ streaming potential measurements, in terms of particle coverage. Such dependencies obtained for various ionic strengths and pH, were successfully interpreted in terms of the 3D electrokinetic model. A universal calibrating graph was produced in this way, enabling one to determine silver monolayer coverage from the measured value of the streaming potential. Our experimental data prove that it is feasible to produce uniform and stable silver particle monolayers of well-controlled coverage and defined electrokinetic properties.

  6. Real-time TIRF observation of vinculin recruitment to stretched α-catenin by AFM.

    Science.gov (United States)

    Maki, Koichiro; Han, Sung-Woong; Hirano, Yoshinori; Yonemura, Shigenobu; Hakoshima, Toshio; Adachi, Taiji

    2018-01-25

    Adherens junctions (AJs) adaptively change their intensities in response to intercellular tension; therefore, they integrate tension generated by individual cells to drive multicellular dynamics, such as morphogenetic change in embryos. Under intercellular tension, α-catenin, which is a component protein of AJs, acts as a mechano-chemical transducer to recruit vinculin to promote actin remodeling. Although in vivo and in vitro studies have suggested that α-catenin-mediated mechanotransduction is a dynamic molecular process, which involves a conformational change of α-catenin under tension to expose a cryptic vinculin binding site, there are no suitable experimental methods to directly explore the process. Therefore, in this study, we developed a novel system by combining atomic force microscopy (AFM) and total internal reflection fluorescence (TIRF). In this system, α-catenin molecules (residues 276-634; the mechano-sensitive M 1 -M 3 domain), modified on coverslips, were stretched by AFM and their recruitment of Alexa-labeled full-length vinculin molecules, dissolved in solution, were observed simultaneously, in real time, using TIRF. We applied a physiologically possible range of tensions and extensions to α-catenin and directly observed its vinculin recruitment. Our new system could be used in the fields of mechanobiology and biophysics to explore functions of proteins under tension by coupling biomechanical and biochemical information.

  7. Tip-enhanced Raman mapping with top-illumination AFM.

    Science.gov (United States)

    Chan, K L Andrew; Kazarian, Sergei G

    2011-04-29

    Tip-enhanced Raman mapping is a powerful, emerging technique that offers rich chemical information and high spatial resolution. Currently, most of the successes in tip-enhanced Raman scattering (TERS) measurements are based on the inverted configuration where tips and laser are approaching the sample from opposite sides. This results in the limitation of measurement for transparent samples only. Several approaches have been developed to obtain tip-enhanced Raman mapping in reflection mode, many of which involve certain customisations of the system. We have demonstrated in this work that it is also possible to obtain TERS nano-images using an upright microscope (top-illumination) with a gold-coated Si atomic force microscope (AFM) cantilever without significant modification to the existing integrated AFM/Raman system. A TERS image of a single-walled carbon nanotube has been achieved with a spatial resolution of ∼ 20-50 nm, demonstrating the potential of this technique for studying non-transparent nanoscale materials.

  8. Tip-enhanced Raman mapping with top-illumination AFM

    Energy Technology Data Exchange (ETDEWEB)

    Chan, K L Andrew; Kazarian, Sergei G, E-mail: s.kazarian@imperial.ac.uk [Department of Chemical Engineering, Imperial College London, SW7 2AZ (United Kingdom)

    2011-04-29

    Tip-enhanced Raman mapping is a powerful, emerging technique that offers rich chemical information and high spatial resolution. Currently, most of the successes in tip-enhanced Raman scattering (TERS) measurements are based on the inverted configuration where tips and laser are approaching the sample from opposite sides. This results in the limitation of measurement for transparent samples only. Several approaches have been developed to obtain tip-enhanced Raman mapping in reflection mode, many of which involve certain customisations of the system. We have demonstrated in this work that it is also possible to obtain TERS nano-images using an upright microscope (top-illumination) with a gold-coated Si atomic force microscope (AFM) cantilever without significant modification to the existing integrated AFM/Raman system. A TERS image of a single-walled carbon nanotube has been achieved with a spatial resolution of {approx} 20-50 nm, demonstrating the potential of this technique for studying non-transparent nanoscale materials.

  9. Tip-enhanced Raman mapping with top-illumination AFM

    International Nuclear Information System (INIS)

    Chan, K L Andrew; Kazarian, Sergei G

    2011-01-01

    Tip-enhanced Raman mapping is a powerful, emerging technique that offers rich chemical information and high spatial resolution. Currently, most of the successes in tip-enhanced Raman scattering (TERS) measurements are based on the inverted configuration where tips and laser are approaching the sample from opposite sides. This results in the limitation of measurement for transparent samples only. Several approaches have been developed to obtain tip-enhanced Raman mapping in reflection mode, many of which involve certain customisations of the system. We have demonstrated in this work that it is also possible to obtain TERS nano-images using an upright microscope (top-illumination) with a gold-coated Si atomic force microscope (AFM) cantilever without significant modification to the existing integrated AFM/Raman system. A TERS image of a single-walled carbon nanotube has been achieved with a spatial resolution of ∼ 20-50 nm, demonstrating the potential of this technique for studying non-transparent nanoscale materials.

  10. Wide range local resistance imaging on fragile materials by conducting probe atomic force microscopy in intermittent contact mode

    Energy Technology Data Exchange (ETDEWEB)

    Vecchiola, Aymeric [Laboratoire de Génie électrique et électronique de Paris (GeePs), UMR 8507 CNRS-CentraleSupélec, Paris-Sud and UPMC Universities, 11 rue Joliot-Curie, Plateau de Moulon, 91192 Gif-sur-Yvette (France); Concept Scientific Instruments, ZA de Courtaboeuf, 2 rue de la Terre de Feu, 91940 Les Ulis (France); Unité Mixte de Physique CNRS-Thales UMR 137, 1 avenue Augustin Fresnel, 91767 Palaiseau (France); Chrétien, Pascal; Schneegans, Olivier; Mencaraglia, Denis; Houzé, Frédéric, E-mail: frederic.houze@geeps.centralesupelec.fr [Laboratoire de Génie électrique et électronique de Paris (GeePs), UMR 8507 CNRS-CentraleSupélec, Paris-Sud and UPMC Universities, 11 rue Joliot-Curie, Plateau de Moulon, 91192 Gif-sur-Yvette (France); Delprat, Sophie [Unité Mixte de Physique CNRS-Thales UMR 137, 1 avenue Augustin Fresnel, 91767 Palaiseau (France); UPMC, Université Paris 06, 4 place Jussieu, 75005 Paris (France); Bouzehouane, Karim; Seneor, Pierre; Mattana, Richard [Unité Mixte de Physique CNRS-Thales UMR 137, 1 avenue Augustin Fresnel, 91767 Palaiseau (France); Tatay, Sergio [Molecular Science Institute, University of Valencia, 46980 Paterna (Spain); Geffroy, Bernard [Lab. Physique des Interfaces et Couches minces (PICM), UMR 7647 CNRS-École polytechnique, 91128 Palaiseau (France); Lab. d' Innovation en Chimie des Surfaces et Nanosciences (LICSEN), NIMBE UMR 3685 CNRS-CEA Saclay, 91191 Gif-sur-Yvette (France); and others

    2016-06-13

    An imaging technique associating a slowly intermittent contact mode of atomic force microscopy (AFM) with a home-made multi-purpose resistance sensing device is presented. It aims at extending the widespread resistance measurements classically operated in contact mode AFM to broaden their application fields to soft materials (molecular electronics, biology) and fragile or weakly anchored nano-objects, for which nanoscale electrical characterization is highly demanded and often proves to be a challenging task in contact mode. Compared with the state of the art concerning less aggressive solutions for AFM electrical imaging, our technique brings a significantly wider range of resistance measurement (over 10 decades) without any manual switching, which is a major advantage for the characterization of materials with large on-sample resistance variations. After describing the basics of the set-up, we report on preliminary investigations focused on academic samples of self-assembled monolayers with various thicknesses as a demonstrator of the imaging capabilities of our instrument, from qualitative and semi-quantitative viewpoints. Then two application examples are presented, regarding an organic photovoltaic thin film and an array of individual vertical carbon nanotubes. Both attest the relevance of the technique for the control and optimization of technological processes.

  11. Phase shifting-based debris effect detection in USV-assisted AFM nanomachining

    Energy Technology Data Exchange (ETDEWEB)

    Shi, Jialin [State Key Laboratory of Robotics, Shenyang Institute of Automation, Chinese Academy of Sciences (CAS), Shenyang, Liaoning 110016 (China); University of the Chinese Academy of Sciences, Beijing 100049 (China); Beijing Advanced Innovation Center for Imaging Technology, Capital Normal University, Beijing 100049 (China); Liu, Lianqing, E-mail: lianqingliu@sia.cn [State Key Laboratory of Robotics, Shenyang Institute of Automation, Chinese Academy of Sciences (CAS), Shenyang, Liaoning 110016 (China); Yu, Peng; Cong, Yang [State Key Laboratory of Robotics, Shenyang Institute of Automation, Chinese Academy of Sciences (CAS), Shenyang, Liaoning 110016 (China); Li, Guangyong [Department of Electrical and Computer Engineering, University of Pittsburgh, Pittsburgh, PA 15213 (United States)

    2017-08-15

    Highlights: • The mechanism of the debris effect on machining depth in force control mode operation is analyzed. • The relationship between phase shifting and pile-up of debris is investigated. • The phase shifting-based method is hardly affected by the pile-up of debris. • Debris effect detection by phase shifting-based method is achived. - Abstract: Atomic force microscopy (AFM) mechanical-based lithography attracts much attention in nanomanufacturing due to its advantages of low cost, high precision and high resolution. However, debris effects during mechanical lithography often lead to an unstable machining process and inaccurate results, which limits further applications of AFM-based lithography. There is a lack of a real-time debris detection approach, which is the prerequisite to eventually eliminating the influence of the debris, and of a method that can solve the above problems well. The ultrasonic vibration (USV)-assisted AFM has the ability to sense the machining depth in real time by detecting the phase shifting of cantilever. However, whether the pile-up of debris affect the phase response of cantilever is still lack of investigation. Therefore, we analyzed the mechanism of the debris effect on force control mode and investigated the relationship between phase shifting and pile-up of debris. Theoretical analysis and experimental results reveal that the pile-up of debris have negligible effect on phase shifting of cantilever. Therefore, the phase shifting-based method can detect the debris effect on machining depth in force control mode of AFM machining.

  12. AFM-based force spectroscopy measurements of mature amyloid fibrils of the peptide glucagon

    DEFF Research Database (Denmark)

    Dong, M. D.; Hovgaard, M. B.; Mamdouh, W.

    2008-01-01

    We report on the mechanical characterization of individual mature amyloid fibrils by atomic force microscopy (AFM) and AFM-based single-molecule force spectroscopy (SMFS). These self-assembling materials, formed from the 29-residue amphiphatic peptide hormone glucagon, were found to display...... a reversible elastic behaviour. Based on AFM morphology and SMFS studies, we suggest that the observed elasticity is due to a force-induced conformational transition which is reversible due to the beta-helical conformation of protofibrils, allowing a high degree of extension. The elastic properties...... of such mature fibrils contribute to their high stability, suggesting that the internal hydrophobic interactions of amyloid fibrils are likely to be of fundamental importance in the assembly of amyloid fibrils and therefore for the understanding of the progression of their associated pathogenic disorders...

  13. Feasibility of multi-walled carbon nanotube probes in AFM anodization lithography

    International Nuclear Information System (INIS)

    Choi, Ji Sun; Bae, Sukjong; Ahn, Sang Jung; Kim, Dal Hyun; Jung, Ki Young; Han, Cheolsu; Chung, Chung Choo; Lee, Haiwon

    2007-01-01

    Multi-walled carbon nanotube (CNT) tips were used in atomic force microscope (AFM) anodization lithography to investigate their advantages over conventional tips. The CNT tip required a larger threshold voltage than the mother silicon tip due to the Schottky barrier at the CNT-Si interface. Current-to-voltage curves distinguished the junction property between CNTs and mother tips. The CNT-platinum tip, which is more conductive than the CNT-silicon tip, showed promising results for AFM anodization lithography. Finally, the nanostructures with high aspect ratio were fabricated using a pulsed bias voltage technique as well as the CNT tip

  14. Investigations of environmental induced effects on AlQ3 thin films by AFM phase imaging

    International Nuclear Information System (INIS)

    Shukla, Vivek Kumar; Kumar, Satyendra

    2007-01-01

    Tris(8-hydroxyquinoline) metal complex (AlQ 3 ) is a widely used light-emitting material in organic light emitting devices (OLEDs). The environmental stability is still a major problem with OLEDs and needs further improvement. In this report, an additional feature of Atomic Force Microscopy (AFM) was exploited with the aim to understand the environmental induced effects and physical phenomenon involved on AlQ 3 thin films. We have used phase imaging to identify the presence of other aggregation phases formed after annealing the thin film in different ambient and after white light exposure. An enhanced photoluminescence intensity is observed for the samples annealed in oxygen near 100 deg. C. The enhanced photoluminescence is understood in terms of formation of a new aggregation phase. The phase change and the fraction of new phase is estimated by phase images taken by atomic force microscopy (AFM). Light induced effects on AlQ 3 films exposed to white light in air and vacuum are characterized by atomic force microscopy (AFM) for surface morphology and phases present. The AFM images indicate enhanced crystallinity for the vacuum exposed samples. The phase with increased lifetime and hence enhanced crystallinity for vacuum exposed films has also been found by time correlated single photon counting (TCSPC) measurements. To the best of our knowledge, this study is applied for the first time on this material with the combination of topography and phase imaging in atomic force microscopy (AFM). The major aim was to take advantage of the additional feature of AFM-mode over the conventionally used

  15. A Study of Moisture Damage in Plastomeric Polymer Modified Asphalt Binder Using Functionalized AFM Tips

    Directory of Open Access Journals (Sweden)

    Rafiqul Tarefder

    2011-12-01

    Full Text Available In this study, moisture damage in plastomeric polymer modified asphalt binder is investigated using Atomic Force Microscopy (AFM with chemically functionalized AFM tips. Four different percentages of plastomeric polymers and two antistripping agents such as Kling Beta and Lime are used to modify a base asphalt binder. Chemical functional groups such as -COOH, -CH3, -NH3, and –OH, that are commonly present in plastomeric polymer modified asphalt system, are used to functionalize the AFM tips. The force distance mode of AFM is used to measure the adhesion forces between a modified asphalt sample surface and the functionalized AFM tips. This enables the measurement of adhesion within an asphalt binder system. It is shown that the adhesion force values in dry sample changed substantially from that in wet conditioned samples. It is evident from this study that plastomeric modification does not help reduce moisture damage in asphalt. The percentage change in adhesion forces due to moisture is about 20 nN for the lime modified samples, and about 50 nN for the Kling Beta modified samples. This indicates that lime is more effective than Kling Beta for reducing moisture damage in plastomeric polymer modified asphalt.

  16. Gum tragacanth stabilized green gold nanoparticles as cargos for Naringin loading: A morphological investigation through AFM.

    Science.gov (United States)

    Rao, Komal; Imran, Muhammad; Jabri, Tooba; Ali, Imdad; Perveen, Samina; Shafiullah; Ahmed, Shakil; Shah, Muhammad Raza

    2017-10-15

    Gold nanoparticles (AuNPs) have attracted greater scientific interests for the construction of drugs loading cargos due to their biocompatibility, safety and facile surface modifications. This study deals with the fabrication of gum tragacanth (GT) green AuNPs as carrier for Naringin, a less water soluble therapeutic molecule. The optimized AuNPs were characterized through UV-vis spectroscopy, FT-IR and atomic force microscope (AFM). Naringin loaded nanoparticles were investigated for their bactericidal potentials using Tetrazolium Microplate assay. Morphological studies conducted via AFM revealed spherical shape for AuNPs with nano-range size and stabilized by GT multi-functional groups. The AuNPs acted as carrier for increased amount of Naringin. Upon loading in AuNPs, Naringin An increased in the bactericidal potentials of Naringin was observed after loading on AuNPs against various tested bacterial strains. This was further authenticated by the surface morphological analysis, showing enhanced membrane destabilizing effects of loaded Naringin. The results suggest that GT stabilized green AuNPs can act as effective delivery vehicles for enhancing bactericidal potentials of Naringin. Copyright © 2017 Elsevier Ltd. All rights reserved.

  17. Restoration of high-resolution AFM images captured with broken probes

    Science.gov (United States)

    Wang, Y. F.; Corrigan, D.; Forman, C.; Jarvis, S.; Kokaram, A.

    2012-03-01

    A type of artefact is induced by damage of the scanning probe when the Atomic Force Microscope (AFM) captures a material surface structure with nanoscale resolution. This artefact has a dramatic form of distortion rather than the traditional blurring artefacts. Practically, it is not easy to prevent the damage of the scanning probe. However, by using natural image deblurring techniques in image processing domain, a comparatively reliable estimation of the real sample surface structure can be generated. This paper introduces a novel Hough Transform technique as well as a Bayesian deblurring algorithm to remove this type of artefact. The deblurring result is successful at removing blur artefacts in the AFM artefact images. And the details of the fibril surface topography are well preserved.

  18. High-resolution AFM structure of DNA G-wires in aqueous solution.

    Science.gov (United States)

    Bose, Krishnashish; Lech, Christopher J; Heddi, Brahim; Phan, Anh Tuân

    2018-05-17

    We investigate the self-assembly of short pieces of the Tetrahymena telomeric DNA sequence d[G 4 T 2 G 4 ] in physiologically relevant aqueous solution using atomic force microscopy (AFM). Wire-like structures (G-wires) of 3.0 nm height with well-defined surface periodic features were observed. Analysis of high-resolution AFM images allowed their classification based on the periodicity of these features. A major species is identified with periodic features of 4.3 nm displaying left-handed ridges or zigzag features on the molecular surface. A minor species shows primarily left-handed periodic features of 2.2 nm. In addition to 4.3 and 2.2 nm ridges, background features with periodicity of 0.9 nm are also observed. Using molecular modeling and simulation, we identify a molecular structure that can explain both the periodicity and handedness of the major G-wire species. Our results demonstrate the potential structural diversity of G-wire formation and provide valuable insight into the structure of higher-order intermolecular G-quadruplexes. Our results also demonstrate how AFM can be combined with simulation to gain insight into biomolecular structure.

  19. Electrical characterisation of semiconductor structures using AFM techniques

    International Nuclear Information System (INIS)

    Kovac jr, J.; Kovac, J.; Hotovy, J.; Novotny, I.; Skriniarova, J.; Dutkova, E.; Balaz, P.

    2011-01-01

    The microscopic dimensions appear to be a fundamental limitation to many common measurement techniques. The use of Current-Atomic Force Microscopy (I-AFM) bids a possibility to acquire topography image along with the current flow mappings which can be lapped over in resulting image as presented in this paper. A current distribution on the ZnO surface of p-Si/n- ZnO diode structure with CdS or ZnS nanocrystalline quantum dot clusters at the interface has been measured. The resulting images show a conductivity mapping different from topography what induces a conductive channels at the edges of the ZnO grains. We have successfully used I-AFM method where conductive AFM tip is scanning over the surface of the sample to create a topography image along with a current flow mapping of p-Si substrate covered with CdS or ZnS nanocrystalline clusters overlapped by 100 nm thick n-ZnO layer. The measured current mappings of both samples revealed a formation of conductive channels between the clusters of quantum dots when the sample is forward biased. We are able to create 3D topography images of combined with the forward biased current mapping textures which gives complex information about local conductivity and using this method it should be possible to find hidden current leaks in the samples for example defects in most semiconductor materials. A drift current generated in p-n junction was recorded when the sample was reverse biased while the sample has been exposed to light. Possible UV light source should cause a higher reverse current due to high bandgap of ZnS clusters which is a motivation to further research. The devices fabricated from these structures have the potential applications for solar cells or broadband photodetectors. (authors)

  20. Metrological AFMs and its application for versatile nano-dimensional metrology tasks

    Science.gov (United States)

    Dai, Gaoliang; Dziomba, T.; Pohlenz, F.; Danzebrink, H.-U.; Koenders, L.

    2010-08-01

    Traceable calibrations of various micro and nano measurement devices are crucial tasks for ensuring reliable measurements for micro and nanotechnology. Today metrological AFM are widely used for traceable calibrations of nano dimensional standards. In this paper, we introduced the developments of metrological force microscopes at PTB. Of the three metrological AFMs described here, one is capable of measuring in a volume of 25 mm x 25 mm x 5 mm. All instruments feature interferometers and the three-dimensional position measurements are thus directly traceable to the metre definition. Some calibration examples on, for instance, flatness standards, step height standards, one and two dimensional gratings are demonstrated.

  1. Vacuolar structures can be identified by AFM elasticity mapping

    International Nuclear Information System (INIS)

    Riethmueller, Christoph; Schaeffer, Tilman E.; Kienberger, Ferry; Stracke, Werner; Oberleithner, Hans

    2007-01-01

    Fluid-filled organelles like vesicles, endosomes and pinosomes are inevitable parts of cellular signalling and transport. Endothelial cells, building a barrier between blood and tissue, can form vacuolar organelles. These structures are implicated in upregulated fluid transport across the endothelium under inflammatory conditions. Vacuolar organelles have been described by transmission electron microscopy so far. Here, we present a method that images and mechanically characterizes intracellular structures in whole cells by atomic force microscopy (AFM). After crosslinking the cellular proteins with the fixative glutaraldehyde, plasma membrane depressions become observable, which are scattered around the cell nucleus. Nanomechanical analysis identifies them as spots of reduced stiffness. Scanning electron microscopy confirms their pit-like appearance. In addition, fluorescence microscopy detects an analogous pattern of protein-poor spots, thereby confirming mechanical rigidity to arise from crosslinked proteins. This AFM application opens up a mechanical dimension for the investigation of intracellular organelles

  2. Self-assembled silver nanoparticles monolayers on mica-AFM, SEM, and electrokinetic characteristics.

    Science.gov (United States)

    Oćwieja, Magdalena; Morga, Maria; Adamczyk, Zbigniew

    2013-03-01

    A monodisperse silver particle suspension was produced by a chemical reduction method in an aqueous medium using sodium citrate. The average particle size determined by dynamic light scattering (DLS), transmission electron microscopy (TEM), and atomic force microscopy (AFM) was 28.5 nm. The DLS measurements confirmed that the suspension was stable for the ionic strength up to 3 × 10 -2  M NaCl. The electrophoretic mobility measurements revealed that the electrokinetic charge of particles was negative for pH range 3-10, assuming -50  e for pH = 9 and 0.01 M NaCl. Using the suspension, silver particle monolayers on mica modified by poly(allylamine hydrochloride) were produced under diffusion-controlled transport. Monolayer coverage, quantitatively determined by AFM and SEM, was regulated within broad limits by adjusting the nanoparticle deposition time. This allowed one to uniquely express the zeta potential of silver monolayers, determined by the in situ streaming potential measurements, in terms of particle coverage. Such dependencies obtained for various ionic strengths and pH, were successfully interpreted in terms of the 3D electrokinetic model. A universal calibrating graph was produced in this way, enabling one to determine silver monolayer coverage from the measured value of the streaming potential. Our experimental data prove that it is feasible to produce uniform and stable silver particle monolayers of well-controlled coverage and defined electrokinetic properties.

  3. Structural analysis of radiation-induced chromosome aberrations by atomic force microscope (AFM) before and after Giemsa staining

    International Nuclear Information System (INIS)

    Murakami, M.; Kanda, R.; Minamihisamatsu, M.; Hayata, I.

    2003-01-01

    Full text: We have studied structures of chromosome aberration induced by ionizing radiation by an atomic force microscope (AFM). The AFM could visualize the fine structure of chromosomes on Giemsa stained or unstained samples, although it was difficult to visualize unstained chromosomes by light microscope. The height data of chromosomes obtained by AFM provided useful information to describe detailed structure of chromatid gaps induced by heavy ion irradiation. A fibrous structure was observed on the unstained chromosome and these structures were considered to be the 30nm fibers on the chromosome. These types of structures were observed in the gaps as well as on surface of the chromosome. Further more, other types of chromosome aberration induced by ionizing radiation visualized by AFM will be presented

  4. High Throughput Nanofabrication of Silicon Nanowire and Carbon Nanotube Tips on AFM Probes by Stencil-Deposited Catalysts

    DEFF Research Database (Denmark)

    Engstrøm, Daniel Southcott; Savu, Veronica; Zhu, Xueni

    2011-01-01

    scale nanostencil lithography. Individual vertical SiNWs were grown epitaxially by a catalytic vapor−liquid−solid (VLS) process and MWNTs were grown by a plasma-enhanced chemical vapor (PECVD) process on the AFM probes. The AFM probes were tested for imaging micrometers-deep trenches, where...... they demonstrated a significantly better performance than commercial high aspect ratio tips. Our method demonstrates a reliable and cost-efficient route toward wafer scale manufacturing of SiNW and MWNT AFM probes....

  5. Refining the statistical model for quantitative immunostaining of surface-functionalized nanoparticles by AFM.

    Science.gov (United States)

    MacCuspie, Robert I; Gorka, Danielle E

    2013-10-01

    Recently, an atomic force microscopy (AFM)-based approach for quantifying the number of biological molecules conjugated to a nanoparticle surface at low number densities was reported. The number of target molecules conjugated to the analyte nanoparticle can be determined with single nanoparticle fidelity using antibody-mediated self-assembly to decorate the analyte nanoparticles with probe nanoparticles (i.e., quantitative immunostaining). This work refines the statistical models used to quantitatively interpret the observations when AFM is used to image the resulting structures. The refinements add terms to the previous statistical models to account for the physical sizes of the analyte nanoparticles, conjugated molecules, antibodies, and probe nanoparticles. Thus, a more physically realistic statistical computation can be implemented for a given sample of known qualitative composition, using the software scripts provided. Example AFM data sets, using horseradish peroxidase conjugated to gold nanoparticles, are presented to illustrate how to implement this method successfully.

  6. Large area nano-patterning /writing on gold substrate using dip - pen nanolithography (DPN)

    Science.gov (United States)

    Saini, Sudhir Kumar; Vishwakarma, Amit; Agarwal, Pankaj B.; Pesala, Bala; Agarwal, Ajay

    2014-10-01

    Dip Pen Nanolithography (DPN) is utilized to pattern large area (50μmX50μm) gold substrate for application in fabricating Nano-gratings. For Nano-writing 16-MHA ink coated AFM tip was prepared using double dipping procedure. Gold substrate is fabricated on thermally grown SiO2 substrate by depositing ˜5 nm titanium layer followed by ˜30nm gold using DC pulse sputtering. The gratings were designed using period of 800nm and 25% duty cycle. Acquired AFM images indicate that as the AFM tip proceeds for nano-writing, line width decreases from 190nm to 100nm. This occurs probably due to depreciation of 16-MHA molecules in AFM tip as writing proceeds.

  7. AFM measurements of novel solar cells. Studying electronic properties of Si-based radial junctions

    Czech Academy of Sciences Publication Activity Database

    Hývl, Matěj

    -, č. 1 (2014), s. 52-53 ISSN 1439-4243 R&D Projects: GA ČR GA13-25747S; GA ČR GA13-12386S; GA MŠk(CZ) LM2011026 Institutional support: RVO:68378271 Keywords : AFM measurements * conductive cantilever * electronic properties * nanowires * PF TUNA Subject RIV: BM - Solid Matter Physics ; Magnetism http://www.imaging-git.com/science/scanning-probe-microscopy/afm-measurements-novel-solar- cells

  8. Quantitative numerical method for analysing slip traces observed by AFM

    International Nuclear Information System (INIS)

    Veselý, J; Cieslar, M; Coupeau, C; Bonneville, J

    2013-01-01

    Atomic force microscopy (AFM) is used more and more routinely to study, at the nanometre scale, the slip traces produced on the surface of deformed crystalline materials. Taking full advantage of the quantitative height data of the slip traces, which can be extracted from these observations, requires however an adequate and robust processing of the images. In this paper an original method is presented, which allows the fitting of AFM scan-lines with a specific parameterized step function without any averaging treatment of the original data. This yields a quantitative and full description of the changes in step shape along the slip trace. The strength of the proposed method is established on several typical examples met in plasticity by analysing nano-scale structures formed on the sample surface by emerging dislocations. (paper)

  9. Ultra-high aspect ratio replaceable AFM tips using deformation-suppressed focused ion beam milling

    International Nuclear Information System (INIS)

    Savenko, Alexey; Yildiz, Izzet; Petersen, Dirch Hjorth; Bøggild, Peter; Bartenwerfer, Malte; Krohs, Florian; Oliva, Maria; Harzendorf, Torsten

    2013-01-01

    Fabrication of ultra-high aspect ratio exchangeable and customizable tips for atomic force microscopy (AFM) using lateral focused ion beam (FIB) milling is presented. While on-axis FIB milling does allow high aspect ratio (HAR) AFM tips to be defined, lateral milling gives far better flexibility in terms of defining the shape and size of the tip. Due to beam-induced deformation, it has so far not been possible to define HAR structures using lateral FIB milling. In this work we obtain aspect ratios of up to 45, with tip diameters down to 9 nm, by a deformation-suppressing writing strategy. Several FIB milling strategies for obtaining sharper tips are discussed. Finally, assembly of the HAR tips on a custom-designed probe as well as the first AFM scanning is shown. (paper)

  10. Initial Growth Process of Magnetron Sputtering 321 Stainless Steel Films Observed by Afm

    Science.gov (United States)

    Jin, Yongzhong; Wu, Wei; Liu, Dongliang; Chen, Jian; Sun, Yali

    To investigate the initial morphological evolution of 321 stainless steel (SS) films, we examined the effect of sputtering time on the morphology of 321 SS film. In this study, a group of samples were prepared at nine different sputtering times within 20 s using radio-frequency (r.f.) magnetron sputtering and characterized by atomic force microscopy (AFM). Only globular-like grains were formed on mica substrates within 6 s, whose average grain size is ~ 21-44 nm. Meanwhile, few grains with larger size are subject to settle at the defect sites of mica substrates. At 8 s, we found large columnar crystallites with the average grain size of 61 nm. From 10 to 14 s, islands grew continuously and coalesced in order to form an interconnected structure containing irregular channels or grooves, with a depth of ~ 3.5-5 nm. Up to 16 s, a nearly continuous film was formed and some new globular-like grains were again present on the film. Study of the AFM image at 20 s suggests that the watercolor masking method designed by us is an effective method, by which we can prepare thin films with steps for the measurement of the thickness of continuous thin films. It is also found that the coverage rate of films increases with the increase in sputtering time (from 2 to 16 s). On the other hand, the increase in root mean square (RMS) roughness is much more significant from 6 to 10 s, and there is a maximum value, 2.81 nm at 10 s due to more islands during deposition. However, RMS roughness decreases with the decrease in length and width of channels or grooves from 10 to 16 s. Especially, a lower RMS roughness of 0.73 nm occurs at 16 s, because of the continuous film produced with a large coverage rate of 98.43%.

  11. AFM lateral force calibration for an integrated probe using a calibration grating

    International Nuclear Information System (INIS)

    Wang, Huabin; Gee, Michelle L.

    2014-01-01

    Atomic force microscopy (AFM) friction measurements on hard and soft materials remain a challenge due to the difficulties associated with accurately calibrating the cantilever for lateral force measurement. One of the most widely accepted lateral force calibration methods is the wedge method. This method is often used in a simplified format but in so doing sacrifices accuracy. In the present work, we have further developed the wedge method to provide a lateral force calibration method for integrated AFM probes that is easy to use without compromising accuracy and reliability. Raw friction calibration data are collected from a single scan image by continuous ramping of the set point as the facets of a standard grating are scanned. These data are analysed to yield an accurate lateral force conversion/calibration factor that is not influenced by adhesion forces or load deviation. By demonstrating this new calibration method, we illustrate its reliability, speed and ease of execution. This method makes accessible reliable boundary lubrication studies on adhesive and heterogeneous surfaces that require spatial resolution of frictional forces. - Highlights: • We develop a simple and accurate method for lateral force calibration in AFM friction measurements. • We detail the basis of the method and illustrate how to use it and its reliability with example data. • Our method is easy, accurate and accounts for the affects of adhesion on friction measurements. • The method is applicable to integrated probes, as opposed to colloidal probes. • This allows accurate AFM friction measurements on spatially heterogeneous and adhesive surfaces

  12. Temperature-Controlled High-Speed AFM: Real-Time Observation of Ripple Phase Transitions.

    Science.gov (United States)

    Takahashi, Hirohide; Miyagi, Atsushi; Redondo-Morata, Lorena; Scheuring, Simon

    2016-11-01

    With nanometer lateral and Angstrom vertical resolution, atomic force microscopy (AFM) has contributed unique data improving the understanding of lipid bilayers. Lipid bilayers are found in several different temperature-dependent states, termed phases; the main phases are solid and fluid phases. The transition temperature between solid and fluid phases is lipid composition specific. Under certain conditions some lipid bilayers adopt a so-called ripple phase, a structure where solid and fluid phase domains alternate with constant periodicity. Because of its narrow regime of existence and heterogeneity ripple phase and its transition dynamics remain poorly understood. Here, a temperature control device to high-speed atomic force microscopy (HS-AFM) to observe dynamics of phase transition from ripple phase to fluid phase reversibly in real time is developed and integrated. Based on HS-AFM imaging, the phase transition processes from ripple phase to fluid phase and from ripple phase to metastable ripple phase to fluid phase could be reversibly, phenomenologically, and quantitatively studied. The results here show phase transition hysteresis in fast cooling and heating processes, while both melting and condensation occur at 24.15 °C in quasi-steady state situation. A second metastable ripple phase with larger periodicity is formed at the ripple phase to fluid phase transition when the buffer contains Ca 2+ . The presented temperature-controlled HS-AFM is a new unique experimental system to observe dynamics of temperature-sensitive processes at the nanoscopic level. © 2016 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  13. Direct AFM observation of an opening event of a DNA cuboid constructed via a prism structure.

    Science.gov (United States)

    Endo, Masayuki; Hidaka, Kumi; Sugiyama, Hiroshi

    2011-04-07

    A cuboid structure was constructed using a DNA origami design based on a square prism structure. The structure was characterized by atomic force microscopy (AFM) and dynamic light scattering. The real-time opening event of the cuboid was directly observed by high-speed AFM.

  14. AFM of metallic nano-particles and nano-structures in heavily irradiated NaCl

    OpenAIRE

    Gaynutdinov, R; Vainshtein, DI; Hak, SJ; Tolstikhina, A; Den Hartog, HW

    2003-01-01

    AFM investigations are reported for heavily, electron irradiated NaCl crystals in ultra high vacuum (UHV) in the non-contact mode-with an UHV AFM/STM Omicron system. To avoid chemical reactions between the radiolytic Na and oxygen and water, the irradiated samples were cleaved and prepared for the experiments in UHV At the surface of freshly cleaved samples, we have observed sodium nano-precipitates with shapes, which depend on the irradiation dose and the volume fraction of the radiolytic Na...

  15. High precision attachment of silver nanoparticles on AFM tips by dielectrophoresis.

    Science.gov (United States)

    Leiterer, Christian; Wünsche, Erik; Singh, Prabha; Albert, Jens; Köhler, Johann M; Deckert, Volker; Fritzsche, Wolfgang

    2016-05-01

    AFM tips are modified with silver nanoparticles using an AC electrical field. The used technique works with sub-micron precision and also does not require chemical modification of the tip. Based on the electrical parameters applied in the process, particle density and particle position on the apex of the tip can be adjusted. The feasibility of the method is proven by subsequent tip-enhanced Raman spectroscopy (TERS) measurements using the fabricated tips as a measurement probe. Since this modification process itself does not require any lithographic processing, the technique can be easily adapted to modify AFM tips with a variety of nanostructures with pre-defined properties, while being parallelizable for a potential commercial application.

  16. Direct comparison of AFM and SEM measurements on the same set of nanoparticles

    International Nuclear Information System (INIS)

    Delvallée, A; Feltin, N; Ducourtieux, S; Trabelsi, M; Hochepied, J F

    2015-01-01

    This article is the first step in the development of a hybrid metrology combining AFM and SEM techniques for measuring the dimensions of a nanoparticle population in 3D space (X,Y,Z). This method exploits the strengths of each technique on the same set of nanoparticles. AFM is used for measuring the nanoparticle height and the measurements along X and Y axes are deduced from SEM images. A sampling method is proposed in order to obtain the best deposition conditions of SiO 2 and gold nanoparticles on mica or silicon substrates. Only the isolated nanoparticles are taken into account in the histogram of size distribution. Moreover, a semi-automatic Matlab routine has also been developed to process the AFM and SEM images, measure and count the nanoparticles. This routine allows the user to exclusively select the isolated nanoparticles through a control interface. The measurements have been performed on spherical-like nanoparticles to test the method by comparing the results obtained with both techniques. (paper)

  17. Surface electrical properties of stainless steel fibres: An AFM-based study

    International Nuclear Information System (INIS)

    Yin, Jun; D’Haese, Cécile; Nysten, Bernard

    2015-01-01

    Highlights: • Surface electrical conductivity of stainless steel fibre is measured and mapped by CS-AFM. • Surface potential of stainless steel fibre is measured and mapped by KPFM. • Surface electronic properties are governed by the chromium oxide passivation layer. • Electron tunnelling through the passivation layer is the dominant mechanisms for conduction. - Abstract: Atomic force microscopy (AFM) electrical modes were used to study the surface electrical properties of stainless steel fibres. The surface electrical conductivity was studied by current sensing AFM and I–V spectroscopy. Kelvin probe force microscopy was used to measure the surface contact potential. The oxide film, known as passivation layer, covering the fibre surface gives rise to the observation of an apparently semiconducting behaviour. The passivation layer generally exhibits a p-type semiconducting behaviour, which is attributed to the predominant formation of chromium oxide on the surface of the stainless steel fibres. At the nanoscale, different behaviours are observed from points to points, which may be attributed to local variations of the chemical composition and/or thickness of the passivation layer. I–V curves are well fitted with an electron tunnelling model, indicating that electron tunnelling may be the predominant mechanism for electron transport

  18. Understanding the extremely low fracture toughness of freestanding gold thin films by in-situ bulge testing in an AFM

    Energy Technology Data Exchange (ETDEWEB)

    Preiß, Eva I., E-mail: eva.preiss@fau.de; Merle, Benoit; Göken, Mathias

    2017-04-13

    The fracture toughness of freestanding gold films with thicknesses between 60 nm and 320 nm was determined by bulge testing to be around 2 MPa m{sup 1/2}. This surprisingly low value confirms the trend also observed for other metals that thin films exhibit only a fraction of the bulk fracture toughness. In order to understand this behavior, the fracture process of freestanding gold films with a crack introduced by focused ion beam (FIB) milling was observed in-situ in an atomic force microscope (AFM). AFM scans of the crack tip region show stable crack growth mainly along grain boundaries. Plastic deformation is localized in a narrow corridor in front of the crack tip. A large plastic zone, as one would typically expect under plane stress, is not observed. Instead, strong local necking is evidenced. We conclude that the spatial confinement of the plastic deformation is the primary reason for the low fracture toughness of metallic thin films.

  19. Accurate Calibration and Uncertainty Estimation of the Normal Spring Constant of Various AFM Cantilevers

    Directory of Open Access Journals (Sweden)

    Yunpeng Song

    2015-03-01

    Full Text Available Measurement of force on a micro- or nano-Newton scale is important when exploring the mechanical properties of materials in the biophysics and nanomechanical fields. The atomic force microscope (AFM is widely used in microforce measurement. The cantilever probe works as an AFM force sensor, and the spring constant of the cantilever is of great significance to the accuracy of the measurement results. This paper presents a normal spring constant calibration method with the combined use of an electromagnetic balance and a homemade AFM head. When the cantilever presses the balance, its deflection is detected through an optical lever integrated in the AFM head. Meanwhile, the corresponding bending force is recorded by the balance. Then the spring constant can be simply calculated using Hooke’s law. During the calibration, a feedback loop is applied to control the deflection of the cantilever. Errors that may affect the stability of the cantilever could be compensated rapidly. Five types of commercial cantilevers with different shapes, stiffness, and operating modes were chosen to evaluate the performance of our system. Based on the uncertainty analysis, the expanded relative standard uncertainties of the normal spring constant of most measured cantilevers are believed to be better than 2%.

  20. Interactions between chitosan and cells measured by AFM

    Energy Technology Data Exchange (ETDEWEB)

    Hsiao, Sheng-Wen; Thien, Doan Van Hong; Ho, Ming-Hua [Department of Chemical Engineering, National Taiwan University of Science and Technology, Taipei 10617, Taiwan (China); Hsieh, Hsyue-Jen [Department of Chemical Engineering, National Taiwan University, Taipei 10617, Taiwan (China); Li, Chung-Hsing [Division of Orthodontics and Pediatric Dentistry, Department of Dentistry, Tri-Service General Hospital, Taipei, Taiwan (China); Hung, Chang-Hsiang [Department of Dentistry, Kinmen Hospital Department of Health, Taiwan (China); Li, Hsi-Hsin, E-mail: mhho@mail.ntust.edu.t [Deputy Superintendent, Kinmen Hospital Department of Health, Taiwan (China)

    2010-10-01

    Chitosan, a biocompatible material that has been widely used in bone tissue engineering, is believed to have a high affinity to osteoblastic cells. This research is the first to prove this hypothesis. By using atomic force microscopy (AFM) with a chitosan-modified cantilever, quantitative evaluation of the interforce between chitosan and cells was carried out. A chitosan tip functionalized with Arg-Gly-Asp (RGD) was also used to measure the interforce between RGD-chitosan and osteoblastic cells. This research concluded by examining cell adhesion and spreading of chitosan substrates as further characterization of the interactions between cells and chitosan. The force measured by AFM showed that the interforce between chitosan and osteoblasts was the highest (209 nN). The smallest adhesion force (61.8 nN) appeared between chitosan and muscle fibroblasts, which did not demonstrate any osteoblastic properties. This result proved that there was a significant interaction between chitosan and bone cells, and correlated with the observations of cell attachment and spreading. The technique developed in this research directly quantified the adhesion between chitosan and cells. This is the first study to demonstrate that specific interaction exists between chitosan and osteoblasts.

  1. Interactions between chitosan and cells measured by AFM

    International Nuclear Information System (INIS)

    Hsiao, Sheng-Wen; Thien, Doan Van Hong; Ho, Ming-Hua; Hsieh, Hsyue-Jen; Li, Chung-Hsing; Hung, Chang-Hsiang; Li, Hsi-Hsin

    2010-01-01

    Chitosan, a biocompatible material that has been widely used in bone tissue engineering, is believed to have a high affinity to osteoblastic cells. This research is the first to prove this hypothesis. By using atomic force microscopy (AFM) with a chitosan-modified cantilever, quantitative evaluation of the interforce between chitosan and cells was carried out. A chitosan tip functionalized with Arg-Gly-Asp (RGD) was also used to measure the interforce between RGD-chitosan and osteoblastic cells. This research concluded by examining cell adhesion and spreading of chitosan substrates as further characterization of the interactions between cells and chitosan. The force measured by AFM showed that the interforce between chitosan and osteoblasts was the highest (209 nN). The smallest adhesion force (61.8 nN) appeared between chitosan and muscle fibroblasts, which did not demonstrate any osteoblastic properties. This result proved that there was a significant interaction between chitosan and bone cells, and correlated with the observations of cell attachment and spreading. The technique developed in this research directly quantified the adhesion between chitosan and cells. This is the first study to demonstrate that specific interaction exists between chitosan and osteoblasts.

  2. Influence of the surface chemistry on the nanotribological behaviour of (AFM tip/graphite) couples

    International Nuclear Information System (INIS)

    Jradi, Khalil; Schmitt, Marjorie; Bistac, Sophie

    2012-01-01

    The development of the nanotechnology has made essential the knowledge of the tribological behaviour of carbonaceous materials, and more particularly of graphite. Atomic force microscopy (AFM) is thus used to study the friction properties at this nanoscopic scale. In this work, results concerning the friction of AFM tips against graphite pins are presented, with a particular emphasis on the effect of the chemical modification of these tips on the tribological behaviour of graphite.

  3. An improved in situ measurement of offset phase shift towards quantitative damping-measurement with AFM

    International Nuclear Information System (INIS)

    Minary-Jolandan, Majid; Yu Minfeng

    2008-01-01

    An improved approach is introduced in damping measurement with atomic force microscope (AFM) for the in situ measurement of the offset phase shift needed for determining the intrinsic mechanical damping in nanoscale materials. The offset phase shift is defined and measured at a point of zero contact force according to the deflection part of the AFM force plot. It is shown that such defined offset phase shift is independent of the type of sample material, varied from hard to relatively soft materials in this study. This improved approach allows the self-calibrated and quantitative damping measurement with AFM. The ability of dynamic mechanical analysis for the measurement of damping in isolated one-dimensional nanostructures, e.g. individual multiwalled carbon nanotubes, was demonstrated

  4. AFM Structural Characterization of Drinking Water Biofilm ...

    Science.gov (United States)

    Due to the complexity of mixed culture drinking water biofilm, direct visual observation under in situ conditions has been challenging. In this study, atomic force microscopy (AFM) revealed the three dimensional morphology and arrangement of drinking water relevant biofilm in air and aqueous solution. Operating parameters were optimized to improve imaging of structural details for a mature biofilm in liquid. By using a soft cantilever (0.03 N/m) and slow scan rate (0.5 Hz), biofilm and individual bacterial cell’s structural topography were resolved and continuously imaged in liquid without loss of spatial resolution or sample damage. The developed methodology will allow future in situ investigations to temporally monitor mixed culture drinking water biofilm structural changes during disinfection treatments. Due to the complexity of mixed culture drinking water biofilm, direct visual observation under in situ conditions has been challenging. In this study, atomic force microscopy (AFM) revealed the three dimensional morphology and arrangement of drinking water relevant biofilm in air and aqueous solution. Operating parameters were optimized to improve imaging of structural details for a mature biofilm in liquid. By using a soft cantilever (0.03 N/m) and slow scan rate (0.5 Hz), biofilm and individual bacterial cell’s structural topography were resolved and continuously imaged in liquid without loss of spatial resolution or sample damage. The developed methodo

  5. Mechanical design, analysis and testing of a large-range compliant microgripper

    Directory of Open Access Journals (Sweden)

    Y. Liu

    2016-04-01

    Full Text Available This paper presents the mechanical design, analysis, fabrication, and testing procedures of a new large-range microgripper which is based on a flexible hinge structure. The uniqueness of the gripper is that the gripper arms not only provide large gripping range but also deliver approximately rectilinear movement as the displacement in nonworking direction is extremely small. The large gripping range is enabled by a mechanism design based on dual-stage flexure amplifier to magnify the stroke of piezoelectric actuator. The first-stage amplifier is a modified version of the Scott Russell (SR mechanism and the second-stage amplifier contains a parallel mechanism. The displacement amplification ratio of the modified SR mechanism in the gripper has been enlarged to 3.56 times of the conventional design. Analytical static models of the gripper mechanism are developed and validated through finite-element analysis (FEA simulation. Results show that the gripping range is over 720 µm with a resonant frequency of 70.7 Hz and negligible displacement in nonworking direction. The total amplification ratio of the input displacement is 16.13. Moreover, a prototype of the gripper is developed by using aluminium 7075 for experimental testing. Experimental results validate the analytical model and FEA simulation results. The proposed microgripper can be employed in various microassembly applications such as pick-and-place of optical fibre.

  6. Nano-Electrochemistry and Nano-Electrografting with an Original Combined AFM-SECM

    Directory of Open Access Journals (Sweden)

    Ammar Ben Brahim

    2013-05-01

    Full Text Available This study demonstrates the advantages of the combination between atomic force microscopy and scanning electrochemical microscopy. The combined technique can perform nano-electrochemical measurements onto agarose surface and nano-electrografting of non-conducting polymers onto conducting surfaces. This work was achieved by manufacturing an original Atomic Force Microscopy-Scanning ElectroChemical Microscopy (AFM-SECM electrode. The capabilities of the AFM-SECM-electrode were tested with the nano-electrografting of vinylic monomers initiated by aryl diazonium salts. Nano-electrochemical and technical processes were thoroughly described, so as to allow experiments reproducing. A plausible explanation of chemical and electrochemical mechanisms, leading to the nano-grafting process, was reported. This combined technique represents the first step towards improved nano-processes for the nano-electrografting.

  7. Controlled AFM manipulation of small nanoparticles and assembly of hybrid nanostructures

    International Nuclear Information System (INIS)

    Kim, Suenne; Shafiei, Farbod; Ratchford, Daniel; Li Xiaoqin

    2011-01-01

    We demonstrate controlled manipulation of semiconductor and metallic nanoparticles (NPs) with 5-15 nm diameters and assemble these NPs into hybrid structures. The manipulation is accomplished under ambient environment using a commercial atomic force microscope (AFM). There are particular difficulties associated with manipulating NPs this small. In addition to spatial drift, the shape of an asymmetric AFM tip has to be taken into account in order to understand the intended and actual manipulation results. Furthermore, small NPs often attach to the tip via electrostatic interaction and modify the effective tip shape. We suggest a method for detaching the NPs by performing a pseudo-manipulation step. Finally, we show by example the ability to assemble these small NPs into prototypical hybrid nanostructures with well-defined composition and geometry.

  8. Nano-Electrochemistry and Nano-Electrografting with an Original Combined AFM-SECM

    Science.gov (United States)

    Ghorbal, Achraf; Grisotto, Federico; Charlier, Julienne; Palacin, Serge; Goyer, Cédric; Demaille, Christophe; Ben Brahim, Ammar

    2013-01-01

    This study demonstrates the advantages of the combination between atomic force microscopy and scanning electrochemical microscopy. The combined technique can perform nano-electrochemical measurements onto agarose surface and nano-electrografting of non-conducting polymers onto conducting surfaces. This work was achieved by manufacturing an original Atomic Force Microscopy-Scanning ElectroChemical Microscopy (AFM-SECM) electrode. The capabilities of the AFM-SECM-electrode were tested with the nano-electrografting of vinylic monomers initiated by aryl diazonium salts. Nano-electrochemical and technical processes were thoroughly described, so as to allow experiments reproducing. A plausible explanation of chemical and electrochemical mechanisms, leading to the nano-grafting process, was reported. This combined technique represents the first step towards improved nano-processes for the nano-electrografting. PMID:28348337

  9. Hematite/silver nanoparticle bilayers on mica--AFM, SEM and streaming potential studies.

    Science.gov (United States)

    Morga, Maria; Adamczyk, Zbigniew; Oćwieja, Magdalena; Bielańska, Elżbieta

    2014-06-15

    Bilayers of hematite/silver nanoparticles were obtained in the self-assembly process and thoroughly characterized using scanning electron microscopy (SEM), atomic force microscopy (AFM), and in situ streaming potential measurements. The hematite nanoparticles, forming a supporting layer, were 22 nm in diameter, exhibiting an isoelectric point at pH 8.9. The silver nanoparticles, used to obtain an external layer, were 29 nm in diameter, and remained negative within the pH range 3 to 11. In order to investigate the particle deposition, mica sheets were used as a model solid substrate. The coverage of the supporting layer was adjusted by changing the bulk concentration of the hematite suspension and the deposition time. Afterward, silver nanoparticle monolayers of controlled coverage were deposited under the diffusion-controlled transport. The coverage of bilayers was determined by a direct enumeration of deposited particles from SEM micrographs and AFM images. Additionally, the formation of the hematite/silver bilayers was investigated by streaming potential measurements carried out under in situ conditions. The effect of the mica substrate and the coverage of a supporting layer on the zeta potential of bilayers was systematically studied. It was established that for the coverage exceeding 0.20, the zeta potential of bilayers was independent on the substrate and the supporting layer coverage. This behavior was theoretically interpreted in terms of the 3D electrokinetic model. Beside significance for basic sciences, these measurements allowed to develop a robust method of preparing nanoparticle bilayers of controlled properties, having potential applications in catalytic processes. Copyright © 2014 Elsevier Inc. All rights reserved.

  10. Optical Beam Deflection Based AFM with Integrated Hardware and Software Platform for an Undergraduate Engineering Laboratory

    Directory of Open Access Journals (Sweden)

    Siu Hong Loh

    2017-02-01

    Full Text Available Atomic force microscopy (AFM has been used extensively in nanoscience research since its invention. Recently, many teaching laboratories in colleges, undergraduate institutions, and even high schools incorporate AFM as an effective teaching tool for nanoscience education. This paper presents an optical beam deflection (OBD based atomic force microscope, designed specifically for the undergraduate engineering laboratory as a teaching instrument. An electronic module for signal conditioning was built with components that are commonly available in an undergraduate electronic laboratory. In addition to off-the-shelf mechanical parts and optics, the design of custom-built mechanical parts waskept as simple as possible. Hence, the overall cost for the setup is greatly reduced. The AFM controller was developed using National Instruments Educational Laboratory Virtual Instrumentation Suite (NI ELVIS, an integrated hardware and software platform which can be programmed in LabVIEW. A simple yet effective control algorithm for scanning and feedback control was developed. Despite the use of an educational platform and low-cost components from the undergraduate laboratory, the developed AFM is capable of performing imaging in constant-force mode with submicron resolution and at reasonable scanning speed (approximately 18 min per image. Therefore, the AFM is suitable to be used as an educational tool for nanoscience. Moreover, the construction of the system can be a valuable educational experience for electronic and mechanical engineering students.

  11. AFM Nanotools for Surgery of Biological Cells

    Energy Technology Data Exchange (ETDEWEB)

    Beard, J D; Gordeev, S N [Department of Physics, Claverton Down, University of Bath, Bath, BA2 7AY (United Kingdom); Guy, R H, E-mail: jdb28@bath.ac.uk [Department of Pharmacy and Pharmacology, Claverton Down, University of Bath, Bath, BA2 7AY (United Kingdom)

    2011-03-01

    Using a method of electron-beam induced deposition, we have been able to fabricate specialized AFM probes with application as 'nanotools' for the manipulation of biological structures ('nanosurgery'). We describe several such tools, including a 'nanoscalpel', 'nanoneedles' for probing intracellular structures, and a 'nanotome' which can separate surface layers from a biological structure. These applications are demonstrated by performing nanomanipulation on corneocyte cells from the outer layer of human skin.

  12. Modular design of AFM probe with sputtered silicon tip

    DEFF Research Database (Denmark)

    Rasmussen, Peter; Thaysen, Jacob; Bouwstra, Siebe

    2001-01-01

    of the thin films constituting the cantilever. The AFM probe has an integrated tip made of a thick sputtered silicon layer, which is deposited after the probe has been defined and just before the cantilevers are released. The tips are so-called rocket tips made by reactive ion etching. We present probes...

  13. Using XAFS, EDAX and AFM in comparative study of various natural and synthetic emeralds

    International Nuclear Information System (INIS)

    Parikh, P.; Saini, N.L.; Dalela, S.; Bhardwaj, D.M.; Fernandes, S.; Gupta, R.P.; Garg, K.B.

    2003-01-01

    We have performed XAFS, EDAX and AFM studies on some natural and synthetic emeralds. While the XAFS results yield information on changes in the valence of the Cr ion and the n-n distance the AFM is used to determine the areal atomic density on surface of the crystals. It is a pilot study to explore if the three techniques can offer a possible way of distinguishing between the natural and synthetic emeralds and the results are promising

  14. Simulation of CNT-AFM tip based on finite element analysis for targeted probe of the biological cell

    Energy Technology Data Exchange (ETDEWEB)

    Yousefi, Amin Termeh, E-mail: at.tyousefi@gmail.com; Miyake, Mikio, E-mail: miyakejaist@gmail.com; Ikeda, Shoichiro, E-mail: sho16.ikeda@gmail.com [ChECA IKohza, Dept. Environmental & Green Technology (EGT), Malaysia, Japan International Institute of Technology (MJIIT), University Technology Malaysia - UTM, Kualalumpur (Malaysia); Mahmood, Mohamad Rusop, E-mail: nano@uitm.gmail.com [NANO-SciTech Centre, Institute of Science, Universiti Teknologi MARA (UiTM), Shah Alam, Selangor (Malaysia)

    2016-07-06

    Carbon nanotubes (CNTs) are potentially ideal tips for atomic force microscopy (AFM) due to the robust mechanical properties, nano scale diameter and also their ability to be functionalized by chemical and biological components at the tip ends. This contribution develops the idea of using CNTs as an AFM tip in computational analysis of the biological cell’s. Finite element analysis employed for each section and displacement of the nodes located in the contact area was monitored by using an output database (ODB). This reliable integration of CNT-AFM tip process provides a new class of high performance nanoprobes for single biological cell analysis.

  15. Fracture Mechanics Testing of Titanium 6AL-4V in AF-M315E

    Science.gov (United States)

    Sampson, J. W.; Martinez, J.; McLean, C.

    2016-01-01

    The Green Propellant Infusion Mission (GPIM) will demonstrate the performance of AF-M315E monopropellant on orbit. Flight certification requires a safe-life analysis of the titanium alloy fuel tank to ensure inherent processing flaws will not cause failure during the design life of the tank. Material property inputs for this analysis require testing to determine the stress intensity factor for environment-assisted cracking (KEAC) of Ti 6Al-4V in combination with the AF-M315E monopropellant. Testing of single-edge notched, or SE(B), specimens representing the bulk tank membrane and weld material were performed in accordance with ASTM E1681. Specimens with fatigue pre-cracks were loaded into test fixtures so that the crack tips were exposed to AF-M315E at 50 C for a duration of 1,000 hours. Specimens that did not fail during exposure were opened to inspect the crack surfaces for evidence of crack growth. The threshold stress intensity value, KEAC, is the highest applied stress intensity that produced neither a failure of the specimen during the exposure nor showed evidence of crack growth. The threshold stress intensity factor for environment-assisted cracking of the Ti 6Al-4V forged tank material was found to be at least 22 ksivin and at least 31 ksivin for the weld material when exposed to AF-M315E monopropellant.

  16. Joint Research on Scatterometry and AFM Wafer Metrology

    OpenAIRE

    Bodermann, B.; Buhr, E.; Danzebrink, H.U.; Bär, M.; Scholze, F.; Krumrey, M.; Wurm, M.; Klapetek, P.; Hansen, P.E.; Korpelainen, V.; Van Veghel, M.; Yacoot, A.; Siitonen, S.; El Gawhary, O.; Burger, S.

    2011-01-01

    Supported by the European Commission and EURAMET, a consortium of 10 participants from national metrology institutes, universities and companies has started a joint research project with the aim of overcoming current challenges in optical scatterometry for traceable linewidth metrology. Both experimental and modelling methods will be enhanced and different methods will be compared with each other and with specially adapted atomic force microscopy (AFM) and scanning electron microscopy (SEM) m...

  17. Cuticle scale measurement of animal fibers by SEM and AFM

    CSIR Research Space (South Africa)

    Notayi, M

    2013-12-01

    Full Text Available -1 51st Microscopy Society of Southern Africa (MSSA) Conference, Farm Inn, Pretoria, 3-6 December 2013 Cuticle scale measurement of animal fibers by SEM and AFM Notayi M, Engelbrechts JAA, Lee ME, Goosen WE, Hunter L and Botha AF Abstract...

  18. Crystallization of Probucol in Nanoparticles Revealed by AFM Analysis in Aqueous Solution.

    Science.gov (United States)

    Egami, Kiichi; Higashi, Kenjirou; Yamamoto, Keiji; Moribe, Kunikazu

    2015-08-03

    The crystallization behavior of a pharmaceutical drug in nanoparticles was directly evaluated by atomic force microscopy (AFM) force curve measurements in aqueous solution. A ternary spray-dried sample (SPD) was prepared by spray drying the organic solvent containing probucol (PBC), hypromellose (HPMC), and sodium dodecyl sulfate (SDS). The amorphization of PBC in the ternary SPD was confirmed by powder X-ray diffraction (PXRD) and solid-state 13C NMR measurements. A nanosuspension containing quite small particles of 25 nm in size was successfully prepared immediately after dispersion of the ternary SPD into water. Furthermore, solution-state 1H NMR measurements revealed that a portion of HPMC coexisted with PBC as a mixed state in the freshly prepared nanosuspension particles. After storing the nanosuspension at 25 °C, a gradual increase in the size of the nanoparticles was observed, and the particle size changed to 93.9 nm after 7 days. AFM enabled the direct observation of the morphology and agglomeration behavior of the nanoparticles in water. Moreover, AFM force-distance curves were changed from (I) to (IV), depending on the storage period, as follows: (I) complete indentation within an applied force of 1 nN, (II) complete indentation with an applied force of 1-5 nN, (III) partial indentation with an applied force of 5 nN, and (IV) nearly no indentation with an applied force of 5 nN. This stiffness increase of the nanoparticles was attributed to gradual changes in the molecular state of PBC from the amorphous to the crystal state. Solid-state 13C NMR measurements of the freeze-dried samples demonstrated the presence of metastable PBC Form II crystals in the stored nanosuspension, strongly supporting the AFM results.

  19. Soft colloidal probes for AFM force measurements between water droplets in oil

    KAUST Repository

    Vakarelski, Ivan Uriev; Li, Erqiang; Thoroddsen, Sigurdur T

    2014-01-01

    Here we introduce an extension of the atomic force microscopy (AFM) colloidal probe technique, as a simple and reliable experimental approach to measure the interaction forces between small water droplets (~80-160. μm) dispersed in oil. Small water droplets are formed by capillary breakup of a microscale water jet in air, which is forced out of a fine capillary nozzle, and deposited on a superhydrophobic substrate immersed in tetradecane oil medium. In these conditions the water droplets are very loosely attached to the superhydrophobic substrate and are easily picked up with a hydrophobic AFM cantilever to form a soft colloidal probe. Sample force measurements are conducted to demonstrate the capability of the technique.

  20. Soft colloidal probes for AFM force measurements between water droplets in oil

    KAUST Repository

    Vakarelski, Ivan Uriev

    2014-11-01

    Here we introduce an extension of the atomic force microscopy (AFM) colloidal probe technique, as a simple and reliable experimental approach to measure the interaction forces between small water droplets (~80-160. μm) dispersed in oil. Small water droplets are formed by capillary breakup of a microscale water jet in air, which is forced out of a fine capillary nozzle, and deposited on a superhydrophobic substrate immersed in tetradecane oil medium. In these conditions the water droplets are very loosely attached to the superhydrophobic substrate and are easily picked up with a hydrophobic AFM cantilever to form a soft colloidal probe. Sample force measurements are conducted to demonstrate the capability of the technique.

  1. Characterization of deep nanoscale surface trenches with AFM using thin carbon nanotube probes in amplitude-modulation and frequency-force-modulation modes

    International Nuclear Information System (INIS)

    Solares, Santiago D

    2008-01-01

    The characterization of deep surface trenches with atomic force microscopy (AFM) presents significant challenges due to the sharp step edges that disturb the instrument and prevent it from faithfully reproducing the sample topography. Previous authors have developed AFM methodologies to successfully characterize semiconductor surface trenches with dimensions on the order of tens of nanometers. However, the study of imaging fidelity for features with dimensions smaller than 10 nm has not yet received sufficient attention. Such a study is necessary because small features in some cases lead to apparently high-quality images that are distorted due to tip and sample mechanical deformation. This paper presents multi-scale simulations, illustrating common artifacts affecting images of nanoscale trenches taken with fine carbon nanotube probes within amplitude-modulation and frequency-force-modulation AFM (AM-AFM and FFM-AFM, respectively). It also describes a methodology combining FFM-AFM with a step-in/step-out algorithm analogous to that developed by other groups for larger trenches, which can eliminate the observed artifacts. Finally, an overview of the AFM simulation methods is provided. These methods, based on atomistic and continuum simulation, have been previously used to study a variety of samples including silicon surfaces, carbon nanotubes and biomolecules

  2. tRNA conjugation with chitosan nanoparticles: An AFM imaging study.

    Science.gov (United States)

    Agudelo, D; Kreplak, L; Tajmir-Riahi, H A

    2016-04-01

    The conjugation of tRNA with chitosan nanoparticles of different sizes 15,100 and 200 kDa was investigated in aqueous solution using multiple spectroscopic methods and atomic force microscopy (AFM). Structural analysis showed that chitosan binds tRNA via G-C and A-U base pairs as well as backbone PO2 group, through electrostatic, hydrophilic and H-bonding contacts with overall binding constants of KCh-15-tRNA=4.1 (±0.60)×10(3)M(-1), KCh-100-tRNA=5.7 (±0.8)×10(3)M(-1) and KCh-200-tRNA=1.2 (±0.3)×10(4)M(-1). As chitosan size increases more stable polymer-tRNA conjugate is formed. AFM images showed major tRNA aggregation and particle formation occurred as chitosan concentration increased. Even though chitosan induced major biopolymer structural changes, tRNA remains in A-family structure. Copyright © 2015 Elsevier B.V. All rights reserved.

  3. EUV blank defect and particle inspection with high throughput immersion AFM with 1nm 3D resolution

    NARCIS (Netherlands)

    Es, M.H. van; Sadeghian Marnani, H.

    2016-01-01

    Inspection of EUV mask substrates and blanks is demanding. We envision this is a good target application for massively parallel Atomic Force Microscopy (AFM). We envision to do a full surface characterization of EUV masks with AFM enabling 1nm true 3D resolution over the entire surface. The limiting

  4. Transmitted wavefront testing with large dynamic range based on computer-aided deflectometry

    Science.gov (United States)

    Wang, Daodang; Xu, Ping; Gong, Zhidong; Xie, Zhongmin; Liang, Rongguang; Xu, Xinke; Kong, Ming; Zhao, Jun

    2018-06-01

    The transmitted wavefront testing technique is demanded for the performance evaluation of transmission optics and transparent glass, in which the achievable dynamic range is a key issue. A computer-aided deflectometric testing method with fringe projection is proposed for the accurate testing of transmitted wavefronts with a large dynamic range. Ray tracing of the modeled testing system is carried out to achieve the virtual ‘null’ testing of transmitted wavefront aberrations. The ray aberration is obtained from the ray tracing result and measured slope, with which the test wavefront aberration can be reconstructed. To eliminate testing system modeling errors, a system geometry calibration based on computer-aided reverse optimization is applied to realize accurate testing. Both numerical simulation and experiments have been carried out to demonstrate the feasibility and high accuracy of the proposed testing method. The proposed testing method can achieve a large dynamic range compared with the interferometric method, providing a simple, low-cost and accurate way for the testing of transmitted wavefronts from various kinds of optics and a large amount of industrial transmission elements.

  5. Measurement of nanosize etched pits in SiO2 optical fiber conduit using AFM

    International Nuclear Information System (INIS)

    Espinosa, G.; Golzarri, J.I.; Vazquez, C.; Fragoso, R.

    2003-01-01

    Fission fragment tracks from 252 Cf have been observed in SiO 2 optical fiber, using an atomic force microscope (AFM), after a very short chemical etching in hydrofluoric acid solution at normal temperature. The nuclear track starting and evolution process is followed by the AFM direct measurements on the material surface and beyond a fine layer of the surface material. The images of the scanned cones were determined observing the two predominant energies from 252 Cf fission fragments and the development of the tracks in the 150 μm diameter optical fiber conduit

  6. A novel dog-bone oscillating AFM probe with thermal actuation and piezoresistive detection.

    Science.gov (United States)

    Xiong, Zhuang; Mairiaux, Estelle; Walter, Benjamin; Faucher, Marc; Buchaillot, Lionel; Legrand, Bernard

    2014-10-31

    In order to effectively increase the resonance frequency and the quality factor of atomic force microscope (AFM) probes, a novel oscillating probe based on a dog-bone shaped MEMS resonator was conceived, designed, fabricated and evaluated. The novel probe with 400 μm in length, 100 μm in width and 5 μm in thickness was enabled to feature MHz resonance frequencies with integrated thermal actuation and piezoresistive detection. Standard silicon micromachining was employed. Both electrical and optical measurements were carried out in air. The resonance frequency and the quality factor of the novel probe were measured to be 5.4 MHz and 4000 respectively, which are much higher than those (about several hundreds of kHz) of commonly used cantilever probes. The probe was mounted onto a commercial AFM set-up through a dedicated probe-holder and circuit board. Topographic images of patterned resist samples were obtained. It is expected that the resonance frequency and the measurement bandwidth of such probes will be further increased by a proper downscaling, thus leading to a significant increase in the scanning speed capability of AFM instruments.

  7. On the nonlinear dynamics of trolling-mode AFM: Analytical solution using multiple time scales method

    Science.gov (United States)

    Sajjadi, Mohammadreza; Pishkenari, Hossein Nejat; Vossoughi, Gholamreza

    2018-06-01

    Trolling mode atomic force microscopy (TR-AFM) has resolved many imaging problems by a considerable reduction of the liquid-resonator interaction forces in liquid environments. The present study develops a nonlinear model of the meniscus force exerted to the nanoneedle of TR-AFM and presents an analytical solution to the distributed-parameter model of TR-AFM resonator utilizing multiple time scales (MTS) method. Based on the developed analytical solution, the frequency-response curves of the resonator operation in air and liquid (for different penetration length of the nanoneedle) are obtained. The closed-form analytical solution and the frequency-response curves are validated by the comparison with both the finite element solution of the main partial differential equations and the experimental observations. The effect of excitation angle of the resonator on horizontal oscillation of the probe tip and the effect of different parameters on the frequency-response of the system are investigated.

  8. Assembly of live micro-organisms on microstructured PDMS stamps by convective/capillary deposition for AFM bio-experiments

    International Nuclear Information System (INIS)

    Dague, E; Jauvert, E; Laplatine, L; Thibault, C; Viallet, B; Ressier, L

    2011-01-01

    Immobilization of live micro-organisms on solid substrates is an important prerequisite for atomic force microscopy (AFM) bio-experiments. The method employed must immobilize the cells firmly enough to enable them to withstand the lateral friction forces exerted by the tip during scanning but without denaturing the cell interface. In this work, a generic method for the assembly of living cells on specific areas of substrates is proposed. It consists in assembling the living cells within the patterns of microstructured, functionalized poly-dimethylsiloxane (PDMS) stamps using convective/capillary deposition. This versatile approach is validated by applying it to two systems of foremost importance in biotechnology and medicine: Saccharomyces cerevisiae yeasts and Aspergillus fumigatus fungal spores. We show that this method allows multiplexing AFM nanomechanical measurements by force spectroscopy on S. cerevisiae yeasts and high-resolution AFM imaging of germinated Aspergillus conidia in buffer medium. These two examples clearly demonstrate the immense potential of micro-organism assembly on functionalized, microstructured PDMS stamps by convective/capillary deposition for performing rigorous AFM bio-experiments on living cells.

  9. Electrochemical Single‐Molecule AFM of the Redox Metalloenzyme Copper Nitrite Reductase in Action

    DEFF Research Database (Denmark)

    Hao, Xian; Zhang, Jingdong; Christensen, Hans Erik Mølager

    2012-01-01

    We studied the electrochemical behavior of the redox metalloenzyme copper nitrite reductase (CNiR, Achromobacter xylosoxidans) immobilized on a Au(111)‐electrode surface modified by a self‐assembled cysteamine molecular monolayer (SAM) using a combination of cyclic voltammetry and electrochemically......‐controlled atomic force microscopy (in situ AFM). The enzyme showed no voltammetric signals in the absence of nitrite substrate, whereas a strong reductive electrocatalytic signal appeared in the presence of nitrite. Such a pattern is common in protein film and monolayer voltammetry and points to conformational...... in the presence of nitrite. No change in size was observed in the absence of nitrite over the same potential range. The enzyme size variation is suggested to offer clues to the broadly observed substrate triggering in metalloenzyme monolayer voltammetry....

  10. Statistical flaw strength distributions for glass fibres: Correlation between bundle test and AFM-derived flaw size density functions

    International Nuclear Information System (INIS)

    Foray, G.; Descamps-Mandine, A.; R’Mili, M.; Lamon, J.

    2012-01-01

    The present paper investigates glass fibre flaw size distributions. Two commercial fibre grades (HP and HD) mainly used in cement-based composite reinforcement were studied. Glass fibre fractography is a difficult and time consuming exercise, and thus is seldom carried out. An approach based on tensile tests on multifilament bundles and examination of the fibre surface by atomic force microscopy (AFM) was used. Bundles of more than 500 single filaments each were tested. Thus a statistically significant database of failure data was built up for the HP and HD glass fibres. Gaussian flaw distributions were derived from the filament tensile strength data or extracted from the AFM images. The two distributions were compared. Defect sizes computed from raw AFM images agreed reasonably well with those derived from tensile strength data. Finally, the pertinence of a Gaussian distribution was discussed. The alternative Pareto distribution provided a fair approximation when dealing with AFM flaw size.

  11. Mechanism of high-resolution STM/AFM imaging with functionalized tips

    Czech Academy of Sciences Publication Activity Database

    Hapala, Prokop; Kichin, G.; Wagner, C.; Tautz, F.S.; Temirov, R.; Jelínek, Pavel

    2014-01-01

    Roč. 90, č. 8 (2014), "085421-1"-"085421-9" ISSN 1098-0121 R&D Projects: GA ČR(CZ) GC14-16963J Grant - others:AVČR(CZ) M100101207 Institutional support: RVO:68378271 Keywords : AFM * STM * high resolution Subject RIV: BM - Solid Matter Physics ; Magnetism Impact factor: 3.736, year: 2014

  12. Accessible light detection and ranging: estimating large tree density for habitat identification

    Science.gov (United States)

    Heather A. Kramer; Brandon M. Collins; Claire V. Gallagher; John Keane; Scott L. Stephens; Maggi Kelly

    2016-01-01

    Large trees are important to a wide variety of wildlife, including many species of conservation concern, such as the California spotted owl (Strix occidentalis occidentalis). Light detection and ranging (LiDAR) has been successfully utilized to identify the density of large-diameter trees, either by segmenting the LiDAR point cloud into...

  13. Measuring the force of single protein molecule detachment from surfaces with AFM.

    Science.gov (United States)

    Tsapikouni, Theodora S; Missirlis, Yannis F

    2010-01-01

    Atomic force microscopy (AFM) was used to measure the non-specific detachment force of single fibrinogen molecules from glass surfaces. The identification of single unbinding events was based on the characteristics of the parabolic curves, recorded during the stretching of protein molecules. Fibrinogen molecules were covalently bound to Si(3)N(4) AFM tips, previously modified with 3-aminopropyl-dimethyl-ethoxysilane, through a homobifunctional poly(ethylene glycol) linker bearing two hydroxysulfosuccinimide esters. The most probable detachment force was found to be 210 pN, when the tip was retracting with a velocity of 1400 nm/s, while the distribution of the detachment distances indicated that the fibrinogen chain can be elongated beyond the length of the physical conformation before detachment. The dependence of the most probable detachment force on the loading rate was examined and the dynamics of fibrinogen binding to the surface were found amenable to the simple expression of the Bell-Evans theory. The theory's expansion, however, by incorporating the concept of the rupture of parallel residue-surface bonds could only describe the detachment of fibrinogen for a small number of such bonds. Finally, the mathematical expression of the Worm-Like Chain model was used to fit the stretching curves before rupture and two interpretations are suggested for the description of the AFM curves with multiple detachment events.

  14. High-resolution AFM topographs of Rubrivivax gelatinosus light-harvesting complex LH2

    Science.gov (United States)

    Scheuring, Simon; Reiss-Husson, Francoise; Engel, Andreas; Rigaud, Jean-Louis; Ranck, Jean-Luc

    2001-01-01

    Light-harvesting complexes 2 (LH2) are the accessory antenna proteins in the bacterial photosynthetic apparatus and are built up of αβ-heterodimers containing three bacteriochlorophylls and one carotenoid each. We have used atomic force microscopy (AFM) to investigate reconstituted LH2 from Rubrivivax gelatinosus, which has a C-terminal hydrophobic extension of 21 amino acids on the α-subunit. High-resolution topographs revealed a nonameric organization of the regularly packed cylindrical complexes incorporated into the membrane in both orientations. Native LH2 showed one surface which protruded by ∼6 Å and one that protruded by ∼14 Å from the membrane. Topographs of samples reconstituted with thermolysin-digested LH2 revealed a height reduction of the strongly protruding surface to ∼9 Å, and a change of its surface appearance. These results suggested that the α-subunit of R.gelatinosus comprises a single transmembrane helix and an extrinsic C-terminus, and allowed the periplasmic surface to be assigned. Occasionally, large rings (∼120 Å diameter) surrounded by LH2 rings were observed. Their diameter and appearance suggest the large rings to be LH1 complexes. PMID:11406579

  15. Binding behavior of CRP and anti-CRP antibody analyzed with SPR and AFM measurement

    International Nuclear Information System (INIS)

    Lee, Soo-Keun; Kim, Hyun-Chul; Cho, Sang-Joon; Jeong, Sang Won; Jeon, Won Bae

    2008-01-01

    Atomic force microscope (AFM) was exploited to take picture of the molecular topology of C-reactive protein (CRP) in phosphate-buffered saline (PBS) solution. An explicit molecular image of CRP demonstrated a pentagonal structure composed of five subunits. Dimensions of the doughnut-shaped CRP molecule measured by AFM were about 25 nm in outside diameter and 10 nm in central pore diameter, and the height of CRP molecule was about 4 nm which was comparable to the value determined by X-ray crystallography. Bis(N-succinimido)-11,11'-dithiobis (undecyl succinate) (DSNHS) was synthesized for use as a linker for immobilizing anti-CRP antibody (anti-CRP) onto the gold surface of a surface plasmon resonance (SPR) sensor chip. DSNHS formed self-assembled monolayer (SAM) on the gold surface. By use of an AFM tip, a pattern of ditch was engraved within the SAM of DSNHS, and anti-CRP was immobilized on the engraved SAM through replacement of N-hydroxysuccinimide group on the outside surface of DSNHS by the amine group of anti-CRP. Formation of CRP/anti-CRP complex on the gold surface of SPR sensor chip was clearly demonstrated by measuring SPR angle shift. A consecutive series of SAM, SAM/anti-CRP, and SAM/anti-CRP/CRP complexes was generated on a SPR sensor chip, and the changes in depth of the ditch were monitored by taking AFM images of the complexes. Comparative analysis of the depth differences indicates that binding of CRP to anti-CRP occurs in a planar mode

  16. Characteristics of a single-channel superconducting flux flow transistor fabricated by an AFM modification technique

    Energy Technology Data Exchange (ETDEWEB)

    Ko, Seokcheol [Jeonnam Regional Innovation Agency, 1000 Namak-Ri, Samhyang-Myun, Muan-Gun, Jeollanam-Do 534-700 (Korea, Republic of)], E-mail: suntrac@jina.re.kr; Kim, Seong-Jong [Mokpo Maritime University, Chukkyo-Dong, Mokpo City, Cheonnam 530-729 (Korea, Republic of)

    2007-11-01

    The demand for high performance, integrity, and miniaturization in the area of electronic and mechanic devices has drawn interest in the fabrication of nanostructures. However, it is difficult to fabricate the channel with nano-scale using a conventional photography techniques. AFM anodization technique is a maskless process and effective method to overcome the difficulty in fabricating a nano-scale channel. In this paper, we first present a new fabrication of a single-channel SFFT using a selective oxidation process induced by an AFM probe. The modified channel was investigated by electron probe microanalyzer (EPMA) to find the compositional variation of the transformed region. In order to confirm the operation of a single-channel SFFT, we measured the voltage-current characteristics at the temperature of liquid nitrogen by an I-V automatic measurement system. Our results indicate that the single-channel SFFT having effect as a weak link is effectively fabricated by an AFM lithography process.

  17. Characteristics of a single-channel superconducting flux flow transistor fabricated by an AFM modification technique

    International Nuclear Information System (INIS)

    Ko, Seokcheol; Kim, Seong-Jong

    2007-01-01

    The demand for high performance, integrity, and miniaturization in the area of electronic and mechanic devices has drawn interest in the fabrication of nanostructures. However, it is difficult to fabricate the channel with nano-scale using a conventional photography techniques. AFM anodization technique is a maskless process and effective method to overcome the difficulty in fabricating a nano-scale channel. In this paper, we first present a new fabrication of a single-channel SFFT using a selective oxidation process induced by an AFM probe. The modified channel was investigated by electron probe microanalyzer (EPMA) to find the compositional variation of the transformed region. In order to confirm the operation of a single-channel SFFT, we measured the voltage-current characteristics at the temperature of liquid nitrogen by an I-V automatic measurement system. Our results indicate that the single-channel SFFT having effect as a weak link is effectively fabricated by an AFM lithography process

  18. Towards easy and reliable AFM tip shape determination using blind tip reconstruction

    International Nuclear Information System (INIS)

    Flater, Erin E.; Zacharakis-Jutz, George E.; Dumba, Braulio G.; White, Isaac A.; Clifford, Charles A.

    2014-01-01

    Quantitative determination of the geometry of an atomic force microscope (AFM) probe tip is critical for robust measurements of the nanoscale properties of surfaces, including accurate measurement of sample features and quantification of tribological characteristics. Blind tip reconstruction, which determines tip shape from an AFM image scan without knowledge of tip or sample shape, was established most notably by Villarrubia [J. Res. Natl. Inst. Stand. Tech. 102 (1997)] and has been further developed since that time. Nevertheless, the implementation of blind tip reconstruction for the general user to produce reliable and consistent estimates of tip shape has been hindered due to ambiguity about how to choose the key input parameters, such as tip matrix size and threshold value, which strongly impact the results of the tip reconstruction. These key parameters are investigated here via Villarrubia's blind tip reconstruction algorithms in which we have added the capability for users to systematically vary the key tip reconstruction parameters, evaluate the set of possible tip reconstructions, and determine the optimal tip reconstruction for a given sample. We demonstrate the capabilities of these algorithms through analysis of a set of simulated AFM images and provide practical guidelines for users of the blind tip reconstruction method. We present a reliable method to choose the threshold parameter corresponding to an optimal reconstructed tip shape for a given image. Specifically, we show that the trend in how the reconstructed tip shape varies with threshold number is so regular that the optimal, or Goldilocks, threshold value corresponds with the peak in the derivative of the RMS difference with respect to the zero threshold curve vs. threshold number. - Highlights: • Blind tip reconstruction algorithms have been implemented and augmented to determine the optimal input parameters. • We demonstrate the capabilities of the algorithms using a simulated AFM

  19. The substructure of immunoglobulin G resolved to 25 kDa using amplitude modulation AFM in air

    International Nuclear Information System (INIS)

    Thomson, Neil H.

    2005-01-01

    Amplitude modulation (or tapping-mode) atomic force microscopy (AM AFM or TM AFM) in air can reveal sub-molecular details of isolated multi-subunit proteins, such as immunoglobulin G (IgG) antibodies, on atomically flat support surfaces such as mica [A. San Paulo, R. Garcia, Biophys. J. 78(3) (2000) 1599]. This is achieved by controlling the microscope imaging parameters (e.g. cantilever drive frequency and set-point amplitude) to keep the AFM tip predominantly in the attractive force regime. Under these conditions, the 50 kDa F c and F ab subunits can be resolved when the molecule has the appropriate orientation on the surface. The presence of a water layer on hydrophilic mica is an important factor affecting imaging contrast, a consequence of capillary neck formation between tip and surface [L. Zitzler, S. Herminghaus, F. Mugele, Phys. Rev. B 66(15) (2002) 155436]. Desiccation of samples to remove surface bound water layers can yield reproducible imaging of the IgG substructure [N.H. Thomson, J. Microsc. (Oxford) 217(3) (2004) 193]. This approach has also given higher resolution than previously achieved, down to about 25 kDa, and these data are detailed here. These subdomains are formed as two immunoglobulin folds from the light and heavy peptide chains of the IgG crossover. This result has been validated by comparing the AFM images with X-ray crystallography data from the protein data bank. These data show that the AFM can obtain 25 kDa resolution on isolated protein molecules with commercially available silicon tips, but, as expected for a local probe technique, resolution is highly dependent on the macromolecular orientation on the support surface

  20. Enhancing local absorption within a gold nano-sphere on a dielectric surface under an AFM probe

    International Nuclear Information System (INIS)

    Talebi Moghaddam, Sina; Ertürk, Hakan; Mengüç, M. Pınar

    2016-01-01

    This study considers enhancing localized absorption by a gold nanoparticle (NP) placed over a substrate where an atomic force microscope (AFM) tip is in close proximity of the particle. The gold NP and AFM tip are interacting with a surface evanescent wave, resulting a near-field coupling between the tip and NP and consequently enhances the absorption. This concept can be used for selective heating of NPs placed over a surface that enables precise manufacturing at nanometer scales. Different tip positions are considered to identify the optimal tip location and the corresponding enhancement limits. The effects of these interactions on the absorption profiles of dielectric core/gold shell NPs are also studied. It is observed that using core–shell nanoparticles with a dielectric core leads to further enhancement of the absorption efficiency and a more uniform distribution of absorption over the shell. Discrete dipole approximation coupled with surface interactions (DDA-SI) is employed throughout the study, and it is vectorized to improve its computational efficiency. - Highlights: • Plasmonic coupling between solid or core-shell nanoparticles, dielectric surface and Si AFM tip is investigated for achieving localized heating for nano-manufacturing. • Absorption efficiency enhancement limits for core-shell and solid nanoparticles are identified using an AFM tip for surface evanescent wave heating. • The effect of tip location, relative to surface wave direction is outlined, identifying optimal locations, and heat absorption distribution over core-shell and solid nanoparticles. • While using a Si AFM tip enhances absorption, using a dielectric core result in further enhancement in absorption with a more uniform distribution. • DDA-SI-v developed by vectorizing the formulations of DDA-SI for improved computational efficiency.

  1. Resonance frequencies of AFM cantilevers in contact with a surface

    Energy Technology Data Exchange (ETDEWEB)

    Verbiest, G.J., E-mail: Verbiest@physik.rwth-aachen.de [JARA-FIT and II. Institute of Physics, RWTH Aachen University, 52074 Aachen (Germany); Rost, M.J., E-mail: Rost@physics.leidenuniv.nl [Huygens-Kamerlingh Onnes Laboratory, Leiden University, P.O. Box 9504, 2300 RA Leiden (Netherlands)

    2016-12-15

    To make the forces in an Atomic Force Microscope that operates in a dynamic mode with one or multiple vibrations applied to the cantilever, quantitative, one needs to relate a change in resonance frequency of the cantilever to a specific tip–sample interaction. Due to the time dependence of the force between the tip and sample caused by the vibrations, this task is not only difficult, but in fact only possible to solve for certain limiting cases, if one follows common theoretical approaches with a Taylor expansion around the deflection point. Here, we present an analytical method for calculating the resonance frequencies of the cantilever that is valid for any tip–sample interaction. Instead of linearizing the tip–sample interaction locally, we calculate an averaged, weighted linearization taking into account all positions of the tip while vibrating. Our method bridges, therefore, the difficult gap between a free oscillating cantilever and a cantilever that is pushed infinitely hard into contact with a surface, which describes a clamped-pinned boundary condition. For a correct description of the cantilever dynamics, we take into account both the tip mass and the tip moment of inertia. Applying our model, we show that it is possible to calculate the modal response of a cantilever as a function of the tip–sample interaction strength. Based on these modal vibration characteristics, we show that the higher resonance frequencies of a cantilever are completely insensitive to the strength of the tip–sample interaction. - Highlights: • A method to calculate the resonances of AFM cantilevers under any force is proposed. • The analytical model is based on Euler-beam theory. • The shift in resonance frequency due to forces decrease with increasing mode number. • The proposed method enables quantitative ultrasound AFM experiments. • Our results explain also the applicability of the higher modes in SubSurface-AFM.

  2. Application of AFM in microscopy and fabrication of micro/nanostructures

    Czech Academy of Sciences Publication Activity Database

    Lopour, F.; Kalousek, R.; Škoda, D.; Spousta, J.; Matějka, František; Šikola, T.

    2002-01-01

    Roč. 34, č. 1 (2002), s. 352 - 355 ISSN 0142-2421 R&D Projects: GA MŠk ME 334; GA MŠk ME 480 Institutional research plan: CEZ:AV0Z2065902 Keywords : AFM fabrication * local anodic oxidation * oxide nanostructures Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering Impact factor: 1.071, year: 2002

  3. Flat ended steel wires, backscattering targets for calibrating over a large dynamic range

    NARCIS (Netherlands)

    Lubbers, Jaap; Graaff, Reindert

    2006-01-01

    A series of flat ended stainless steel wires was constructed and experimentally evaluated as point targets giving a calibrated backscattering over a large range (up to 72 dB) for ultrasound frequencies in the range 2 to 10 MHz. Over a range of 36 dB, theory was strictly followed (within 1 dB),

  4. Force and Compliance Measurements on Living Cells Using Atomic Force Microscopy (AFM

    Directory of Open Access Journals (Sweden)

    Wojcikiewicz Ewa P.

    2004-01-01

    Full Text Available We describe the use of atomic force microscopy (AFM in studies of cell adhesion and cell compliance. Our studies use the interaction between leukocyte function associated antigen-1 (LFA-1/intercellular adhesion molecule-1 (ICAM-1 as a model system. The forces required to unbind a single LFA-1/ICAM-1 bond were measured at different loading rates. This data was used to determine the dynamic strength of the LFA-1/ICAM-1 complex and characterize the activation potential that this complex overcomes during its breakage. Force measurements acquired at the multiple- bond level provided insight about the mechanism of cell adhesion. In addition, the AFM was used as a microindenter to determine the mechanical properties of cells. The applications of these methods are described using data from a previous study.

  5. Carbon nanotube/carbon nanotube composite AFM probes prepared using ion flux molding

    Science.gov (United States)

    Chesmore, Grace; Roque, Carrollyn; Barber, Richard

    The performance of carbon nanotube-carbon nanotube composite (CNT/CNT composite) atomic force microscopy (AFM) probes is compared to that of conventional Si probes in AFM tapping mode. The ion flux molding (IFM) process, aiming an ion beam at the CNT probe, aligns the tip to a desired angle. The result is a relatively rigid tip that is oriented to offset the cantilever angle. Scans using these probes reveal an improvement in image accuracy over conventional tips, while allowing higher aspect ratio imaging of 3D surface features. Furthermore, the lifetimes of CNT-CNT composite tips are observed to be longer than both conventional tips and those claimed for other CNT technologies. Novel applications include the imaging of embiid silk. Supported by the Clare Boothe Luce Research Scholars Award and Carbon Design Innovations.

  6. Magnetocaloric effect and negative thermal expansion in hexagonal Fe doped MnNiGe compounds with a magnetoelastic AFM-FM-like transition.

    Science.gov (United States)

    Xu, Kun; Li, Zhe; Liu, Enke; Zhou, Haichun; Zhang, Yuanlei; Jing, Chao

    2017-01-30

    We report a detailed study of two successive first-order transitions, including a martensitic transition (MT) and an antiferromagnetic (AFM)-ferromagnetic (FM)-like transition, in Mn 1-x Fe x NiGe (x = 0, 0.06, 0.11) alloys by X-ray diffraction, differential scanning calorimetry, magnetization and linear thermal expansion measurements. Such an AFM-FM-like transition occurring in the martensitic state has seldom been observed in the M(T) curves. The results of Arrott plot and linear relationship of the critical temperature with M 2 provide explicit evidence of its first-order magnetoelastic nature. On the other hand, their performances as magnetocaloric and negative thermal expansion materials were characterized. The isothermal entropy change for a field change of 30 kOe reaches an impressive value of -25.8 J/kg K at 203 K for x = 0.11 compared to the other two samples. It demonstrates that the magneto-responsive ability has been significantly promoted since an appropriate amount of Fe doping can break the local Ni-6Mn AFM configuration. Moreover, the Fe-doped samples reveal both the giant negative thermal expansion and near-zero thermal expansion for different temperature ranges. For instance, the average thermal expansion coefficient ā of x = 0.06 reaches -60.7 × 10 -6 /K over T = 231-338 K and 0.6 × 10 -6 /K over T = 175-231 K during cooling.

  7. A surface evolution scheme to identify nanoscale intrinsic geometry from AFM experimental data

    International Nuclear Information System (INIS)

    Jang, Hong-Lae; Kim, Hyun-Seok; Park, Youmie; Cho, Seonho

    2013-01-01

    The geometrical properties of metallic nanoparticles such as the size and morphology have significant impacts on the structure and stability of the adsorbed biological entities as well as the nanoscale structural performances. To identify the nanoscale intrinsic geometry from the height images by atomic force microscopy (AFM), we developed a curvature-dependent evolution scheme that can eliminate the noise and smoothen the surfaces. The principal curvatures are computed directly from the first and second derivatives of the discrete AFM height data. The principal curvatures and directions correspond to the eigenvalues and eigenvectors of shape operator matrix, respectively. The evolution equation using the principal curvature flows smoothens the images in the corresponding principal directions. For an idealized model, κ 2 flow successfully identifies the major valley lines to represent the boundary of nanoparticles without referring to the phase information, whereas the mean curvature flow eliminates all the minor ones leaving only the major feature of the boundary. To demonstrate the capability of noise removal, smoothing surfaces, the identification of ridge and valley lines, and the extraction of intrinsic geometry, the developed numerical scheme is applied to real AFM data that include the silver nanoparticles of 24 nm diameter and the gold nanoparticles of 33–56 nm diameters

  8. Association between Urinary Aflatoxin (AFM1 and Dietary Intake among Adults in Hulu Langat District, Selangor, Malaysia

    Directory of Open Access Journals (Sweden)

    Siti Husna Sulaiman

    2018-04-01

    Full Text Available Aflatoxin is a food contaminant and its exposure through the diet is frequent and ubiquitous. A long-term dietary aflatoxin exposure has been linked to the development of liver cancer in populations with high prevalence of aflatoxin contamination in foods. Therefore, this study was conducted to identify the association between urinary aflatoxin M1 (AFM1, a biomarker of aflatoxin exposure, with the dietary intake among adults in Hulu Langat district, Selangor, Malaysia. Certain food products have higher potential for aflatoxin contamination and these were listed in a Food Frequency Questionnaire, which was given to all study participants. This allowed us to record consumption rates for each food product listed. Concomitantly, urine samples were collected, from adults in selected areas in Hulu Langat district, for the measurement of AFM1 levels using an ELISA kit. Of the 444 urine samples collected and tested, 199 were positive for AFM1, with 37 of them exceeding the limit of detection (LOD of 0.64 ng/mL. Cereal products showed the highest consumption level among all food groups, with an average intake of 512.54 g per day. Chi-square analysis showed that consumption of eggs (X2 = 4.77, p = 0.03 and dairy products (X2 = 19.36, p < 0.01 had significant associations with urinary AFM1 but both food groups were having a phi and Cramer’s V value that less than 0.3, which indicated that the association between these food groups’ consumption and AFM1 level in urine was weak.

  9. AFM Surface Roughness and Topography Analysis of Lithium Disilicate Glass Ceramic

    Directory of Open Access Journals (Sweden)

    M. Pantić

    2015-12-01

    Full Text Available The aim of this study is presenting AFM analysis of surface roughness of Lithium disilicate glass ceramic (IPS e.max CAD under different finishing procedure (techniques: polishing, glazing and grinding. Lithium disilicate glass ceramics is all-ceramic dental system which is characterized by high aesthetic quality and it can be freely said that properties of material provide all prosthetic requirements: function, biocompatibility and aesthetic. Experimental tests of surface roughness were investigated on 4 samples with dimensions: 18 mm length, 14 mm width and 12 mm height. Contact surfaces of three samples were treated with different finishing procedure (polishing, glazing and grinding, and the contact surface of the raw material is investigated as a fourth sample. Experimental measurements were done using the Atomic Force Microscopy (AFM of NT-MDT manufacturers, in the contact mode. All obtained results of different prepared samples are presented in the form of specific roughness parameters (Rа, Rz, Rmax, Rq and 3D surface topography.

  10. Design of a micro-cartridge system for the robotic assembly of exchangeable AFM-probe tips

    DEFF Research Database (Denmark)

    Bartenwerfer, Malte; Eichhorn, Volkmar; Fatikow, Sergej

    2013-01-01

    demand an even higher lateral resolution of the measurements. The atomic force microscope (AFM) is a common tool for this characterization and a standard instrument for all kinds of research and development disciplines. However, the characterization of three dimensional high-aspect ratio and sidewall...... structures remains a hardly accomplishable task. Novel exchangeable and customizable scanning probe tips, so-called NanoBits, can be attached to standard AFM cantilevers offering unprecedented freedom in adapting the shape and size of the tips to the surface topology of the specific application. The ultimate...

  11. A Novel Dog-Bone Oscillating AFM Probe with Thermal Actuation and Piezoresistive Detection †

    Science.gov (United States)

    Xiong, Zhuang; Mairiaux, Estelle; Walter, Benjamin; Faucher, Marc; Buchaillot, Lionel; Legrand, Bernard

    2014-01-01

    In order to effectively increase the resonance frequency and the quality factor of atomic force microscope (AFM) probes, a novel oscillating probe based on a dog-bone shaped MEMS resonator was conceived, designed, fabricated and evaluated. The novel probe with 400 μm in length, 100 μm in width and 5 μm in thickness was enabled to feature MHz resonance frequencies with integrated thermal actuation and piezoresistive detection. Standard silicon micromachining was employed. Both electrical and optical measurements were carried out in air. The resonance frequency and the quality factor of the novel probe were measured to be 5.4 MHz and 4000 respectively, which are much higher than those (about several hundreds of kHz) of commonly used cantilever probes. The probe was mounted onto a commercial AFM set-up through a dedicated probe-holder and circuit board. Topographic images of patterned resist samples were obtained. It is expected that the resonance frequency and the measurement bandwidth of such probes will be further increased by a proper downscaling, thus leading to a significant increase in the scanning speed capability of AFM instruments. PMID:25365463

  12. Single molecule transcription profiling with AFM

    International Nuclear Information System (INIS)

    Reed, Jason; Mishra, Bud; Pittenger, Bede; Magonov, Sergei; Troke, Joshua; Teitell, Michael A; Gimzewski, James K

    2007-01-01

    Established techniques for global gene expression profiling, such as microarrays, face fundamental sensitivity constraints. Due to greatly increasing interest in examining minute samples from micro-dissected tissues, including single cells, unorthodox approaches, including molecular nanotechnologies, are being explored in this application. Here, we examine the use of single molecule, ordered restriction mapping, combined with AFM, to measure gene transcription levels from very low abundance samples. We frame the problem mathematically, using coding theory, and present an analysis of the critical error sources that may serve as a guide to designing future studies. We follow with experiments detailing the construction of high density, single molecule, ordered restriction maps from plasmids and from cDNA molecules, using two different enzymes, a result not previously reported. We discuss these results in the context of our calculations

  13. Neural network approximation of tip-abrasion effects in AFM imaging

    International Nuclear Information System (INIS)

    Bakucz, Peter; Dziomba, Thorsten; Koenders, Ludger; Krüger-Sehm, Rolf; Yacoot, Andrew

    2008-01-01

    The abrasion (wear) of tips used in scanning force microscopy (SFM) directly influences SFM image quality and is therefore of great relevance to quantitative SFM measurements. The increasing implementation of automated SFM measurement schemes has become a strong driving force for increasing efforts towards the prediction of tip wear, as it needs to be ensured that the probe is exchanged before a level of tip wear is reached that adversely affects the measurement quality. In this paper, we describe the identification of tip abrasion in a system of SFM measurements. We attempt to model the tip-abrasion process as a concatenation of a mapping from the measured AFM data to a regression vector and a nonlinear mapping from the regressor space to the output space. The mapping is formed as a basis function expansion. Feedforward neural networks are used to approximate this mapping. The one-hidden layer network gave a good quality of fit for the training and test sets for the tip-abrasion system. We illustrate our method with AFM measurements of both fine periodic structures and randomly oriented sharp features and compare our neural network results with those obtained using other methods

  14. Neural network approximation of tip-abrasion effects in AFM imaging

    Science.gov (United States)

    Bakucz, Peter; Yacoot, Andrew; Dziomba, Thorsten; Koenders, Ludger; Krüger-Sehm, Rolf

    2008-06-01

    The abrasion (wear) of tips used in scanning force microscopy (SFM) directly influences SFM image quality and is therefore of great relevance to quantitative SFM measurements. The increasing implementation of automated SFM measurement schemes has become a strong driving force for increasing efforts towards the prediction of tip wear, as it needs to be ensured that the probe is exchanged before a level of tip wear is reached that adversely affects the measurement quality. In this paper, we describe the identification of tip abrasion in a system of SFM measurements. We attempt to model the tip-abrasion process as a concatenation of a mapping from the measured AFM data to a regression vector and a nonlinear mapping from the regressor space to the output space. The mapping is formed as a basis function expansion. Feedforward neural networks are used to approximate this mapping. The one-hidden layer network gave a good quality of fit for the training and test sets for the tip-abrasion system. We illustrate our method with AFM measurements of both fine periodic structures and randomly oriented sharp features and compare our neural network results with those obtained using other methods.

  15. AFM picking-up manipulation of the metaphase chromosome fragment by using the tweezers-type probe

    International Nuclear Information System (INIS)

    Yamanaka, Keiichiro; Saito, Masato; Shichiri, Motoharu; Sugiyama, Sigeru; Takamura, Yuzuru; Hashiguchi, Gen; Tamiya, Eiichi

    2008-01-01

    We have studied the development of a new procedure based on atomic force microscopy (AFM) for the analysis of metaphase chromosome. The aim of this study was to obtain detailed information about the specific locations of genes on the metaphase chromosome. In this research, we performed the manipulation of the metaphase chromosome by using novel AFM probes to obtain chromosome fragments of a smaller size than the ones obtained using the conventional methods, such as glass microneedles. We could pick up the fragment of the metaphase chromosome dissected by the knife-edged probe by using our tweezers-type probe

  16. Large dynamic range pressure sensor based on two semicircle-holes microstructured fiber.

    Science.gov (United States)

    Liu, Zhengyong; Htein, Lin; Lee, Kang-Kuen; Lau, Kin-Tak; Tam, Hwa-Yaw

    2018-01-08

    This paper presents a sensitive and large dynamic range pressure sensor based on a novel birefringence microstructured optical fiber (MOF) deployed in a Sagnac interferometer configuration. The MOF has two large semicircle holes in the cladding and a rectangular strut with germanium-doped core in the center. The fiber structure permits surrounding pressure to induce large effective index difference between the two polarized modes. The calculated and measured group birefringence of the fiber are 1.49 × 10 -4 , 1.23 × 10 -4 , respectively, at the wavelength of 1550 nm. Experimental results shown that the pressure sensitivity of the sensor varied from 45,000 pm/MPa to 50,000 pm/MPa, and minimum detectable pressure of 80 Pa and dynamic range of better than 116 dB could be achieved with the novel fiber sensor. The proposed sensor could be used in harsh environment and is an ideal candidate for downhole applications where high pressure measurement at elevated temperature up to 250 °C is needed.

  17. 3-Dimensional atomic scale structure of the ionic liquid-graphite interface elucidated by AM-AFM and quantum chemical simulations

    Science.gov (United States)

    Page, Alister J.; Elbourne, Aaron; Stefanovic, Ryan; Addicoat, Matthew A.; Warr, Gregory G.; Voïtchovsky, Kislon; Atkin, Rob

    2014-06-01

    In situ amplitude modulated atomic force microscopy (AM-AFM) and quantum chemical simulations are used to resolve the structure of the highly ordered pyrolytic graphite (HOPG)-bulk propylammonium nitrate (PAN) interface with resolution comparable with that achieved for frozen ionic liquid (IL) monolayers using STM. This is the first time that (a) molecular resolution images of bulk IL-solid interfaces have been achieved, (b) the lateral structure of the IL graphite interface has been imaged for any IL, (c) AM-AFM has elucidated molecular level structure immersed in a viscous liquid and (d) it has been demonstrated that the IL structure at solid surfaces is a consequence of both thermodynamic and kinetic effects. The lateral structure of the PAN-graphite interface is highly ordered and consists of remarkably well-defined domains of a rhomboidal superstructure composed of propylammonium cations preferentially aligned along two of the three directions in the underlying graphite lattice. The nanostructure is primarily determined by the cation. Van der Waals interactions between the propylammonium chains and the surface mean that the cation is enriched in the surface layer, and is much less mobile than the anion. The presence of a heterogeneous lateral structure at an ionic liquid-solid interface has wide ranging ramifications for ionic liquid applications, including lubrication, capacitive charge storage and electrodeposition.In situ amplitude modulated atomic force microscopy (AM-AFM) and quantum chemical simulations are used to resolve the structure of the highly ordered pyrolytic graphite (HOPG)-bulk propylammonium nitrate (PAN) interface with resolution comparable with that achieved for frozen ionic liquid (IL) monolayers using STM. This is the first time that (a) molecular resolution images of bulk IL-solid interfaces have been achieved, (b) the lateral structure of the IL graphite interface has been imaged for any IL, (c) AM-AFM has elucidated molecular level

  18. AFM surface imaging of AISI D2 tool steel machined by the EDM process

    International Nuclear Information System (INIS)

    Guu, Y.H.

    2005-01-01

    The surface morphology, surface roughness and micro-crack of AISI D2 tool steel machined by the electrical discharge machining (EDM) process were analyzed by means of the atomic force microscopy (AFM) technique. Experimental results indicate that the surface texture after EDM is determined by the discharge energy during processing. An excellent machined finish can be obtained by setting the machine parameters at a low pulse energy. The surface roughness and the depth of the micro-cracks were proportional to the power input. Furthermore, the AFM application yielded information about the depth of the micro-cracks is particularly important in the post treatment of AISI D2 tool steel machined by EDM

  19. AFM surface imaging of AISI D2 tool steel machined by the EDM process

    Science.gov (United States)

    Guu, Y. H.

    2005-04-01

    The surface morphology, surface roughness and micro-crack of AISI D2 tool steel machined by the electrical discharge machining (EDM) process were analyzed by means of the atomic force microscopy (AFM) technique. Experimental results indicate that the surface texture after EDM is determined by the discharge energy during processing. An excellent machined finish can be obtained by setting the machine parameters at a low pulse energy. The surface roughness and the depth of the micro-cracks were proportional to the power input. Furthermore, the AFM application yielded information about the depth of the micro-cracks is particularly important in the post treatment of AISI D2 tool steel machined by EDM.

  20. AFM imaging and analysis of local mechanical properties for detection of surface pattern of functional groups

    Energy Technology Data Exchange (ETDEWEB)

    Knotek, Petr, E-mail: petr.knotek@upce.cz [University of Pardubice, Faculty of Chemical Technology, Joint Laboratory of Solid State Chemistry of IMC ASCR and University of Pardubice, Studentska 573, 532 10 Pardubice (Czech Republic); Chanova, Eliska; Rypacek, Frantisek [Institute of Macromolecular Chemistry, Academy of Sciences of the Czech Republic, Heyrovskeho sq. 2, 162 06 Prague (Czech Republic)

    2013-05-01

    In this work we evaluate the applicability of different atomic force microscopy (AFM) modes, such as Phase Shift Imaging, Atomic Force Acoustic Microscopy (AFAM) and Force Spectroscopy, for mapping of the distribution pattern of low-molecular-weight biomimetic groups on polymer biomaterial surfaces. Patterns with either random or clustered spatial distribution of bioactive peptide group derived from fibronectin were prepared by surface deposition of functional block copolymer nano-colloids and grafted with RGDS peptide containing the sequence of amino acids arginine–glycine–aspartic acid–serine (conventionally labeled as RGDS) and carrying biotin as a tag. The biotin-tagged peptides were labeled with 40 nm streptavidin-modified Au nanospheres. The peptide molecules were localized through the detection of bound Au nanospheres by AFM, and thus, the surface distribution of peptides was revealed. AFM techniques capable of monitoring local mechanical properties of the surface were proved to be the most efficient for identification of Au nano-markers. The efficiency was successfully demonstrated on two different patterns, i.e. random and clustered distribution of RGDS peptides on structured surface of the polymer biomaterial. Highlights: ► Bioactive peptides for cell adhesion on PLA-b-PEO biomimetic surface were visualized. ► The biotin-tagged RGDS peptides were labeled with streptavidin-Au nanospheres. ► The RGDS pattern was detected using different atomic force microscopy (AFM) modes. ► Phase Shift Image was proved to be suitable method for studying peptide distribution.

  1. BOREAS AFM-6 Boundary Layer Height Data

    Science.gov (United States)

    Wilczak, James; Hall, Forrest G. (Editor); Newcomer, Jeffrey A. (Editor); Smith, David E. (Technical Monitor)

    2000-01-01

    The Boreal Ecosystem-Atmosphere Study (BOREAS) Airborne Fluxes and Meteorology (AFM)-6 team from National Oceanic and Atmospheric Adminsitration/Environment Technology Laboratory (NOAA/ETL) operated a 915-MHz wind/Radio Acoustic Sounding System (RASS) profiler system in the Southern Study Area (SSA) near the Old Jack Pine (OJP) site. This data set provides boundary layer height information over the site. The data were collected from 21 May 1994 to 20 Sep 1994 and are stored in tabular ASCII files. The boundary layer height data are available from the Earth Observing System Data and Information System (EOSDIS) Oak Ridge National Laboratory (ORNL) Distributed Active Archive Center (DAAC). The data files are available on a CD-ROM (see document number 20010000884).

  2. Surface study of irradiated sapphires from Phrae Province, Thailand using AFM

    Science.gov (United States)

    Monarumit, N.; Jivanantaka, P.; Mogmued, J.; Lhuaamporn, T.; Satitkune, S.

    2017-09-01

    The irradiation is one of the gemstone enhancements for improving the gem quality. Typically, there are many varieties of irradiated gemstones in the gem market such as diamond, topaz, and sapphire. However, it is hard to identify the gemstones before and after irradiation. The aim of this study is to analyze the surface morphology for classifying the pristine and irradiated sapphires using atomic force microscope (AFM). In this study, the sapphire samples were collected from Phrae Province, Thailand. The samples were irradiated by high energy electron beam for a dose of ionizing radiation at 40,000 kGy. As the results, the surface morphology of pristine sapphires shows regular atomic arrangement, whereas, the surface morphology of irradiated sapphires shows the nano-channel observed by the 2D and 3D AFM images. The atomic step height and root mean square roughness have changed after irradiation due to the micro-structural defect on the sapphire surface. Therefore, this study is a frontier application for sapphire identification before and after irradiation.

  3. Automatic determination of the size of elliptical nanoparticles from AFM images

    International Nuclear Information System (INIS)

    Sedlář, Jiří; Zitová, Barbara; Kopeček, Jaromír; Flusser, Jan; Todorciuc, Tatiana; Kratochvílová, Irena

    2013-01-01

    The objective of this work was to develop an accurate method for automatic determination of the size of elliptical nanoparticles from atomic force microscopy (AFM) images that would yield results consistent with results of manual measurements by experts. The proposed method was applied on phenylpyridyldiketopyrrolopyrrole (PPDP), a granular organic material with a wide scale of application and highly sensitive particle-size properties. A PPDP layer consists of similarly sized elliptical particles (c. 100 nm × 50 nm) and its properties can be estimated from the average length and width of the particles. The developed method is based on segmentation of salient particles by the watershed transform and approximation of their shapes by ellipses computed by image moments; it estimates the lengths and widths of the particles by the major and minor axes, respectively, of the corresponding ellipses. Its results proved to be consistent with results of manual measurements by a trained expert. The comparison showed that the developed method could be used in practice for precise automatic measurement of PPDP particles in AFM images

  4. Probing of miniPEGγ-PNA-DNA Hybrid Duplex Stability with AFM Force Spectroscopy.

    Science.gov (United States)

    Dutta, Samrat; Armitage, Bruce A; Lyubchenko, Yuri L

    2016-03-15

    Peptide nucleic acids (PNA) are synthetic polymers, the neutral peptide backbone of which provides elevated stability to PNA-PNA and PNA-DNA hybrid duplexes. It was demonstrated that incorporation of diethylene glycol (miniPEG) at the γ position of the peptide backbone increased the thermal stability of the hybrid duplexes (Sahu, B. et al. J. Org. Chem. 2011, 76, 5614-5627). Here, we applied atomic force microscopy (AFM) based single molecule force spectroscopy and dynamic force spectroscopy (DFS) to test the strength and stability of the hybrid 10 bp duplex. This hybrid duplex consisted of miniPEGγ-PNA and DNA of the same length (γ(MP)PNA-DNA), which we compared to a DNA duplex with a homologous sequence. AFM force spectroscopy data obtained at the same conditions showed that the γ(MP)PNA-DNA hybrid is more stable than the DNA counterpart, 65 ± 15 pN vs 47 ± 15 pN, respectively. The DFS measurements performed in a range of pulling speeds analyzed in the framework of the Bell-Evans approach yielded a dissociation constant, koff ≈ 0.030 ± 0.01 s⁻¹ for γ(MP)PNA-DNA hybrid duplex vs 0.375 ± 0.18 s⁻¹ for the DNA-DNA duplex suggesting that the hybrid duplex is much more stable. Correlating the high affinity of γ(MP)PNA-DNA to slow dissociation kinetics is consistent with prior bulk characterization by surface plasmon resonance. Given the growing interest in γ(MP)PNA as well as other synthetic DNA analogues, the use of single molecule experiments along with computational analysis of force spectroscopy data will provide direct characterization of various modifications as well as higher order structures such as triplexes and quadruplexes.

  5. A comprehensive modeling and vibration analysis of AFM microcantilevers subjected to nonlinear tip-sample interaction forces

    International Nuclear Information System (INIS)

    Eslami, Sohrab; Jalili, Nader

    2012-01-01

    Precise and accurate representation of an Atomic Force Microscopy (AFM) system is essential in studying the effects of boundary interaction forces present between the probe's tip and the sample. In this paper, a comprehensive analytical model for the AFM system utilizing a distributed-parameters based approach is proposed. More specifically, we consider two important attributes of these systems; namely the rotary inertia and shear deformation when compared with the Euler–Bernoulli beam theory. Moreover, a comprehensive nonlinear interaction force is assumed between probe's and sample in order to reveal the response of the system more realistically. This nanoscale interaction force is based on a general form consisting of both attractive and repulsive components as well as a function of the tip-sample distance and the microcantilever's base and sample oscillations. Mechanical properties of the sample could interact with the nanomechanical coupling field between the probe' tip and sample and be implemented in studying the composition information of the sample and the ultra-small features inside it. Therefore, by modulating the dynamics of the AFM system such as the driving amplitude of the microcantilever the procedure for the subsurface imaging is described. The presented approach here could be implemented for designing the AFM probes by examining the tip-sample interaction forces dominant by the van der Waals forces. Several numerical case studies are presented and the force–distance diagram reveals that the proposed nonlinear nanomechanical force along with the distributed-parameters model for the microcantilever is able to fulfill the mechanics of the Lennard–Jones potential. -- Highlights: ► We present a comprehensive distributed-parameters model for AFM microcantilever. ► Assuming a nonlinear and implicit interaction force between tip and sample. ► Timoshenko beam is compared with the Euler–Bernoulli having the same force model. ► Frequency

  6. Confocal Raman and PL, AFM, and X-ray diffraction studies of CdS:O thin films

    International Nuclear Information System (INIS)

    Akinori, Suzuki; Kazuki, Wakita; YongGu, Shim; Nazim, Mamedov; Ayaz, Bayramov; Emil, Huseynov

    2010-01-01

    Full text : CdS has much attention as a window material of thin-film solar cells, for example a CdTe solar cell. In this case, increasing band gap of CdS films leads to rise of conversion efficiency of a solar cell. Recently, it was reported that CdS:O films deposited by rf magnetron sputtering consist of nano-crystals of CdS resulting in increasing the band gap. This work reports confocal Raman and photoluminescence (PL), atomic force microscopy (AFM), and X-ray diffraction studies of CdS:O films deposited by cathode sputtering for formation of nano-crystal of CdS. It was shown that AFM image of CdS:O films annealed at 300, 400 and 500 degrees Celsium. The height of peak and dip on the surface is in the range of 5 and 20 nm in the samples annealed at less than 400 degrees Celsium, while the clear crystalline shape appears in the sample annealed at 500 degrees Celsium. There is also shown X-ray diffraction pattern of CdS:O films. As grown film shows amorphous structure of CdS. On the other hand, the samples annealed at 400 and 500 degrees Celsium display obvious crystalline pattern. The crystal radius of the samples annealed at 300, 400, and 500 degrees Celsium were estimated to be 20, 27, and 37 nm, respectively, according to Scherrers formula. Other results related with the confocal spectroscopy will be also presented.

  7. Observation of Shapiro-steps in AFM-plought micron-size YBCO planar construction

    CSIR Research Space (South Africa)

    Elkaseh, AAO

    2009-01-01

    Full Text Available Using an Atomic Force Microscope (AFM), micron size planar constriction type junctions was successfully ploughed on YBa2Cu3O7-x thin films. The 100 nanometer (nm) thin films are deposited on MgO substrates by an Inverted Cylindrical Magnetron (ICM...

  8. AFM imaging and fractal analysis of surface roughness of AlN epilayers on sapphire substrates

    Energy Technology Data Exchange (ETDEWEB)

    Dallaeva, Dinara, E-mail: dinara.dallaeva@yandex.ru [Brno University of Technology, Faculty of Electrical Engineering and Communication, Physics Department, Technická 8, 616 00 Brno (Czech Republic); Ţălu, Ştefan [Technical University of Cluj-Napoca, Faculty of Mechanical Engineering, Department of AET, Discipline of Descriptive Geometry and Engineering Graphics, 103-105 B-dul Muncii Street, Cluj-Napoca 400641, Cluj (Romania); Stach, Sebastian [University of Silesia, Faculty of Computer Science and Materials Science, Institute of Informatics, Department of Biomedical Computer Systems, ul. Będzińska 39, 41-205 Sosnowiec (Poland); Škarvada, Pavel; Tománek, Pavel; Grmela, Lubomír [Brno University of Technology, Faculty of Electrical Engineering and Communication, Physics Department, Technická 8, 616 00 Brno (Czech Republic)

    2014-09-01

    Graphical abstract: - Highlights: • We determined the complexity of 3D surface roughness of aluminum nitride layers. • We used atomic force microscopy and analyzed their fractal geometry. • We determined the fractal dimension of surface roughness of aluminum nitride layers. • We determined the dependence of layer morphology on substrate temperature. - Abstract: The paper deals with AFM imaging and characterization of 3D surface morphology of aluminum nitride (AlN) epilayers on sapphire substrates prepared by magnetron sputtering. Due to the effect of temperature changes on epilayer's surface during the fabrication, a surface morphology is studied by combination of atomic force microscopy (AFM) and fractal analysis methods. Both methods are useful tools that may assist manufacturers in developing and fabricating AlN thin films with optimal surface characteristics. Furthermore, they provide different yet complementary information to that offered by traditional surface statistical parameters. This combination is used for the first time for measurement on AlN epilayers on sapphire substrates, and provides the overall 3D morphology of the sample surfaces (by AFM imaging), and reveals fractal characteristics in the surface morphology (fractal analysis)

  9. TEM and AFM study of WO3 nanosize growth on α-Al2O3

    International Nuclear Information System (INIS)

    Al-Mohammad, A.

    2007-07-01

    WO 3 thin films have been deposited by thermal evaporation on (0001) and (1012 ) planes of alumina oxide single crystal and annealed either in Oxygen or in air atmosphere. The morphology and crystallographic structure of films (as-deposited and annealed films) have been characterized by Atomic Force Microscope (AFM), transmission electron microscope (TEM), and transmission electron diffraction (TED). During annealing, the films undergo important morphological and structural changes. The annealed films exhibit large grains. These grains have the monoclinic structure in epitaxial orientations. The grains are made of twinned microdomains elongated in the [100] direction resulting of a preferential growth. The microdomains are along the three different directions on the (0001) α-Al 2 O 3 surface and only one direction on the (1012 ) α-Al 2 O 3 one.(author)

  10. Two-dimensional dopant profiling for shallow junctions by TEM and AFM

    International Nuclear Information System (INIS)

    Yoo, K.

    2000-01-01

    The present work concerns the development of the Etch/TEM and Etch/AFM methods to obtain quantitative 2-D dopant profiles for the ultra shallow p-n junctions of the next generation of metal-oxide-semiconductor field effect transistors (MOSFETs). For these methods, thin foil (TEM) or bulk (AFM) cross-sectional specimens were etched using a dopant selective chemical so that local areas of the dopant implanted source/drain (S/D) regions were etched to different depths. The surface topography of the S/D regions was determined from the thickness fringes for the TEM method and by the direct measurement for the AFM method. The local etched depths were converted to etch rates, and these were then converted to corresponding 1-D and 2-D dopant profiles by the experimentally independent etch rate calibration curves. Shallow junction MOSFET samples were designed and fabricated with junction depths 60nm (n + /p), 80nm (n + /p) and 120nm (p + /n) using 0.25μm process technology. A new method using SOG (Spin-on-Glass) contributed to the high quality XTEM thin foil specimens. Controlled stirring of the etchant increased the dopant concentration selectivity and etching consistency. Computer modelling simulated the isotropic etching behaviours, which can introduce the significant error in dopant profiling for shallow and abrupt junction samples. Comprehensive quantitative results enabled the optimum etching time to be determined for the first time. Etch/TEM method gave 1-D dopant profiles that showed good agreement with 1-D Spreading Resistance Probe (SRP) dopant profiles for determining junction depths. 2-D dopant profiles gave L eff , i.e. the shortest lateral distance between the S/D junctions, of major importance for MOSFET performance. Values for L eff of 161, 159 and 123nm were determined from 60, 80 and 120nm junction depth samples respectively, compared with the 215nm MOSFET gate length. The resolution and accuracy of the Etch/TEM method are estimated as 2 and 10nm

  11. Nanotribology at single crystal electrodes: Influence of ionic adsorbates on friction forces studied with AFM

    Energy Technology Data Exchange (ETDEWEB)

    Hausen, Florian; Nielinger, Michael; Ernst, Siegfried [Institut fuer Physikalische und Theoretische Chemie, Universitaet Bonn, Roemerstrasse 164, D-53117 Bonn (Germany); Baltruschat, Helmut [Institut fuer Physikalische und Theoretische Chemie, Universitaet Bonn, Roemerstrasse 164, D-53117 Bonn (Germany)], E-mail: baltruschat@uni-bonn.de

    2008-09-01

    We present friction force measurements on Au(1 1 1) single crystal electrode surfaces performed under electrochemical conditions using an atomic force microscope (AFM). At monoatomic steps friction is increased in both scan directions. In 0.05 M sulfuric acid an increase of friction is observed with the increase of adsorbed sulfate. Friction force increases non-linearly with load. Cu UPD also increases friction in presence of sulfate. However, in presence of 4 x 10{sup -4} M chloride friction is much smaller for all deposited Cu coverages - ranging from a submonolayer up to bulk copper compared to the solution without chloride. After dissolution of bulk copper clusters deposited on Au(1 1 1) we observed an area with higher friction forces due to the formation of an alloy between gold and copper.

  12. Fabrication of tuning-fork based AFM and STM tungsten probe

    KAUST Repository

    Al-Falih, Hisham

    2011-12-01

    We compare the sharpness of tungsten probe tips produced by the single-step and two-step dynamic electrochemical etching processes. A small radius of curvature (RoC) of 25 nm or less was routinely obtained when the two-step electrochemical etching (TEE) process was adopted, while the smallest achievable RoC was ∼10 nm, rendering it suitable for atomic force microscopy (AFM) or scanning tunneling microscopy (STM) applications. © 2011 IEEE.

  13. CHARACTERIZATION OF SURFACE OF THE (010 FACE OF BORAX CRYSTALS USING EX SITU ATOMIC FORCE MICROSCOPY (AFM:

    Directory of Open Access Journals (Sweden)

    Suharso Suharso

    2010-06-01

    Full Text Available The surface topology of borax crystals grown at a relative supersaturation of 0.21 has been investigated using ex situ atomic force microscopy (AFM. It was found that the cleavage of borax crystals along the (010 face planes has features of the cleavage of layered compounds, exhibiting cleavage steps of low heights. The step heights of the cleavage of the (010 face of borax crystal are from one unit cell to three unit cells of this face.   Keywords: AFM, cleavage, borax.

  14. Association between Urinary Aflatoxin (AFM1) and Dietary Intake among Adults in Hulu Langat District, Selangor, Malaysia

    Science.gov (United States)

    Sulaiman, Siti Husna

    2018-01-01

    Aflatoxin is a food contaminant and its exposure through the diet is frequent and ubiquitous. A long-term dietary aflatoxin exposure has been linked to the development of liver cancer in populations with high prevalence of aflatoxin contamination in foods. Therefore, this study was conducted to identify the association between urinary aflatoxin M1 (AFM1), a biomarker of aflatoxin exposure, with the dietary intake among adults in Hulu Langat district, Selangor, Malaysia. Certain food products have higher potential for aflatoxin contamination and these were listed in a Food Frequency Questionnaire, which was given to all study participants. This allowed us to record consumption rates for each food product listed. Concomitantly, urine samples were collected, from adults in selected areas in Hulu Langat district, for the measurement of AFM1 levels using an ELISA kit. Of the 444 urine samples collected and tested, 199 were positive for AFM1, with 37 of them exceeding the limit of detection (LOD) of 0.64 ng/mL. Cereal products showed the highest consumption level among all food groups, with an average intake of 512.54 g per day. Chi-square analysis showed that consumption of eggs (X2 = 4.77, p = 0.03) and dairy products (X2 = 19.36, p food groups were having a phi and Cramer’s V value that less than 0.3, which indicated that the association between these food groups’ consumption and AFM1 level in urine was weak. PMID:29642443

  15. Effects of AFM tip-based direct and vibration assisted scratching methods on nanogrooves fabrication on a polymer resist

    Energy Technology Data Exchange (ETDEWEB)

    Geng, Yanquan [The State Key Laboratory of Robotics and Systems, Robotics Institute, Harbin Institute of Technology, Harbin, Heilongjiang 150080 (China); Center for Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Yan, Yongda, E-mail: yanyongda@hit.edu.cn [The State Key Laboratory of Robotics and Systems, Robotics Institute, Harbin Institute of Technology, Harbin, Heilongjiang 150080 (China); Center for Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China); Zhuang, Yun; Hu, Zhenjiang [Center for Precision Engineering, Harbin Institute of Technology, Harbin, Heilongjiang 150001 (China)

    2015-11-30

    Graphical abstract: - Highlights: • The comparison of three different atomic force microscope (AFM) tip-based material processing techniques to generate nano-grooves on polymethylmethacrylate (PMMA) thin film is presented. • The machined depths of the nano-grooves machined by these three methods are analyzed. • Nano-groove with the machined depth closed to the thickness of the thin-film resist is achieved. - Abstract: This study proposes two atomic force microscope (AFM) tip-based direct nanoscratching methods including single-pass scratching and multi-pass scratching compared with a vibration-assisted scratching method to fabricate nano-grooves on the surface of the polymethylmethacrylate (PMMA) thin-film resist. In order to protect the AFM tip from wearing and optimize the subsequent etching process, the machined depth is expected slightly less than the PMMA thickness to prevent the tip directly contacting with the silicon substrate and obtain better process results. First, single-pass scratching tests are performed on films with different thickness employing varied normal loads. Results show that the machined depths of the grooves cannot be obtained slightly less than the thickness of the film very easily when scratching with single-pass method, 50–120 nm in the present study, which may not be very suitable for the following etching process. Multi-pass and vibration-assisted methods are then utilized to solve this limitation of the machined depth in single-pass process. The machined depths using the multi-pass method are dependent on scratching times and the applied normal loads. Moreover, the depth closed to the thickness of the film can be obtained by enlarging the number of the scratching cycles. However, with a longer scratching time, large tip wear can be found. For vibration assisted method, the machined depths are controlled by the vibration amplitude and the applied normal load. With the vibration in z direction increasing, the machined depth can

  16. Decay characteristics of electronic charged states of Si quantum dots as evaluated by an AFM/Kelvin probe technique

    International Nuclear Information System (INIS)

    Nishitani, Junichiro; Makihara, Katsunori; Ikeda, Mitsuhisa; Murakami, Hideki; Higashi, Seiichiro; Miyazaki, Seiichi

    2006-01-01

    Si quantum dots (Si-QDs) with an areal dot density of 8 x 10 11 cm -2 on SiO 2 have been prepared by the thermal decomposition of monosilane to characterize charged states of Si-QDs using AFM/Kelvin probe force microscopy (KFM). The temporal changes in the surface potential induced by electron charging and discharging at Si-QDs covered with a ∼ 2-nm-thick oxide layer have been measured. In electron charging and discharging at Si-QDs, a Rh-coated AFM tip was electrically biased in the range of - 5 to 5 V and scanned on the sample surface in a tapping mode. The surface potential changes on Si-QDs by electron injection and extraction were observable, while no potential change was detected elsewhere. The surface potential of charged Si-QDs decays with time at rates depending on charge injection conditions. The observed decay characteristics can be interpreted in terms of discharging of stored electrons in Si-QDs due to electron tunneling through the bottom oxide to the substrate and neutralization of stored holes due to recombination with electrons tunneling from the substrates. The defect generation in oxide is likely to be responsible for a fairly slow decay as observed in the case of electron extraction by the tip bias as high as + 4.8 V with respect to p-Si(100)

  17. A study of AFM-based scratch process on polycarbonate surface and grating application

    International Nuclear Information System (INIS)

    Choi, Chul Hyun; Lee, Dong Jin; Sung, Jun-Ho; Lee, Min Woo; Lee, Seung-Gol; Park, Se-Geun; Lee, El-Hang; O, Beom-Hoan

    2010-01-01

    We report on the possibility of applying atomic force microscope (AFM) lithography to draw micro/nano-structures on the surface of a polycarbonate (PC) substrate. We also fabricated a grating structure on the PC surface using the scratch method. An AFM silicon tip coated with a diamond layer was utilized as a cutting tool to scratch the surface of the sample. In order to obtain pattern depth deeper than the control method of interaction force, we used a scanner movement method which the sample scanner moves along the Z-axis. A grating of 100 μm x 150 μm was fabricated by the step and repeat method wherein the sample stage is moved in the direction of the XY-axis. The period and the depth of the grating are 500 and 50 nm, respectively. Light of 632.8 nm wavelength was diffracted on the surface of the PC substrate.

  18. AFM study of excimer laser patterning of block-copolymer: Creation of ordered hierarchical, hybrid, or recessed structures

    International Nuclear Information System (INIS)

    Švanda, Jan; Siegel, Jakub; Švorčík, Vaclav; Lyutakov, Oleksiy

    2016-01-01

    Highlights: • Combination of bottom-up (BCP separation) and top-down (laser patterning) technologies allows obtaining hierarchical structures. • Surface morphologies were determined by the order of patterning steps (laser modification, annealing, surface reconstruction). • Tuning the order of steps enables the reorientation of BCP domain at large scale, fabrication of hierarchical, hybrid or recessed structures. • The obtained structures can find potential applications in nanotechnology, plasmonics, information storage, sensors and smart surfaces. - Abstract: We report fabrication of the varied range of hierarchical structures by combining bottom-up self-assembly of block copolymer poly(styrene-block-vinylpyridine) (PS-b-P4VP) with top-down excimer laser patterning method. Different procedures were tested, where laser treatment was applied before phase separation and after phase separation or phase separation and surface reconstruction. Laser treatment was performed using either polarized laser light with the aim to create periodical pattern on polymer surface or non-polarized light for preferential removing of polystyrene (PS) part from PS-b-P4VP. Additionally, dye was introduced into one part of block copolymer (P4VP) with the aim to modify its response to laser light. Resulting structures were analyzed by XPS, UV–vis and AFM techniques. Application of polarized laser light leads to creation of structures with hierarchical, recessed or hybrid geometries. Non-polarized laser beam allows pronouncing the block copolymer phase separated structure. Tuning the order of steps or individual step conditions enables the efficient reorientation of block-copolymer domain at large scale, fabrication of hierarchical, hybrid or recessed structures. The obtained structures can find potential applications in nanotechnology, photonics, plasmonics, information storage, optical devices, sensors and smart surfaces.

  19. A quantitative AFM analysis of nano-scale surface roughness in various orthodontic brackets.

    Science.gov (United States)

    Lee, Gi-Ja; Park, Ki-Ho; Park, Young-Guk; Park, Hun-Kuk

    2010-10-01

    In orthodontics, the surface roughnesses of orthodontic archwire and brackets affect the effectiveness of arch-guided tooth movement, corrosion behavior, and the aesthetics of orthodontic components. Atomic force microscopy (AFM) measurements were used to provide quantitative information on the surface roughness of the orthodontic material. In this study, the changes in surface roughness of various orthodontic bracket slots before and after sliding movement of archwire in vitro and in vivo were observed through the utilization of AFM. Firstly, we characterized the surface of four types of brackets slots as follows: conventional stainless steel (Succes), conventional ceramic (Perfect), self-ligating stainless steel (Damon) and self-ligating ceramic (Clippy-C) brackets. Succes) and Damon brackets showed relatively smooth surfaces, while Perfect had the roughest surface among the four types of brackets used. Secondly, after in vitro sliding test with beta titanium wire in two conventional brackets (Succes and Perfect), there were significant increases in only stainless steel bracket, Succes. Thirdly, after clinical orthodontic treatment for a maximum of 2 years, the self-ligating stainless steel bracket, Damon, showed a significant increase in surface roughness. But self-ligating ceramic brackets, Clippy-C, represented less significant changes in roughness parameters than self-ligating stainless steel ones. Based on the results of the AFM measurements, it is suggested that the self-ligating ceramic bracket has great possibility to exhibit less friction and better biocompatibility than the other tested brackets. This implies that these bracket slots will aid in the effectiveness of arch-guided tooth movement.

  20. BOREAS AFM-06 Mean Wind Profile Data

    Science.gov (United States)

    Wilczak, James; Hall, Forrest G. (Editor); Newcomer, Jeffrey A. (Editor); Smith, David E. (Technical Monitor)

    2000-01-01

    The Boreal Ecosystem-Atmosphere Study (BOREAS) Airborne Fluxes and Meteorology (AFM)-6 team from the National Oceanic and Atmospheric Administration/Environment Technology Laboratory (NOAA/ETL) operated a 915-MHz wind/Radio Acoustic Sounding System (RASS) profiler system in the Southern Study Area (SSA) near the Old Jack Pine (OJP) tower from 21 May 1994 to 20 Sep 1994. The data set provides wind profiles at 38 heights, containing the variables of wind speed; wind direction; and the u-, v-, and w-components of the total wind. The data are stored in tabular ASCII files. The mean wind profile data are available from the Earth Observing System Data and Information System (EOSDIS) Oak Ridge National Laboratory (ORNL) Distributed Active Archive Center (DAAC). The data files are available on a CD-ROM (see document number 20010000884).

  1. Influence of the tip mass and position on the AFM cantilever dynamics: Coupling between bending, torsion and flexural modes

    Energy Technology Data Exchange (ETDEWEB)

    Mokhtari-Nezhad, F. [Department of Mechanical Engineering, Shahid Bahonar University of Kerman, Kerman (Iran, Islamic Republic of); Saidi, A.R., E-mail: saidi@mail.uk.ac.ir [Department of Mechanical Engineering, Shahid Bahonar University of Kerman, Kerman (Iran, Islamic Republic of); Ziaei-Rad, S. [Department of Mechanical Engineering, Isfahan University of Technology (IUT), Isfahan 84156-83111 (Iran, Islamic Republic of)

    2009-08-15

    The effects of the geometrical asymmetric related to tip position as a concentrated mass, on the sensitivity of all three vibration modes, lateral excitation (LE), torsional resonance (TR) and vertical excitation (VE), of an atomic force microscopy (AFM) microcantilever have been analyzed. The effects of the tip mass and its position are studied to report the novel results to estimating the vibration behavior of AFM such as resonance frequency and amplitude of the microcantilever. In this way, to achieve more accurate results, the coupled motion in all three modes is considered. In particular, it is investigated that performing the coupled motion in analysis of AFM microcantilever is almost necessary. It is shown that the tip mass and its position have significant effects on vibrational responses. The results show that considering the tip mass decreases the resonance frequencies particularly on high-order modes. However, dislocating of tip position has an inverse effect that causes an increase in the resonance frequencies. In addition, it has been shown that the amplitude of the AFM microcantilever is affected by the influences of tip and its position. These effects are caused by the interaction between flexural and torsional motion due to the moment of inertia of the tip mass.

  2. Influence of the tip mass and position on the AFM cantilever dynamics: Coupling between bending, torsion and flexural modes

    International Nuclear Information System (INIS)

    Mokhtari-Nezhad, F.; Saidi, A.R.; Ziaei-Rad, S.

    2009-01-01

    The effects of the geometrical asymmetric related to tip position as a concentrated mass, on the sensitivity of all three vibration modes, lateral excitation (LE), torsional resonance (TR) and vertical excitation (VE), of an atomic force microscopy (AFM) microcantilever have been analyzed. The effects of the tip mass and its position are studied to report the novel results to estimating the vibration behavior of AFM such as resonance frequency and amplitude of the microcantilever. In this way, to achieve more accurate results, the coupled motion in all three modes is considered. In particular, it is investigated that performing the coupled motion in analysis of AFM microcantilever is almost necessary. It is shown that the tip mass and its position have significant effects on vibrational responses. The results show that considering the tip mass decreases the resonance frequencies particularly on high-order modes. However, dislocating of tip position has an inverse effect that causes an increase in the resonance frequencies. In addition, it has been shown that the amplitude of the AFM microcantilever is affected by the influences of tip and its position. These effects are caused by the interaction between flexural and torsional motion due to the moment of inertia of the tip mass.

  3. Investigation of the resistive switching in Ag{sub x}AsS{sub 2} layer by conductive AFM

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, Bo [Department of General and Inorganic Chemistry, Faculty of Chemical Technology, University of Pardubice, Studentska 573, Pardubice, 532 10 Czech Republic (Czech Republic); Kutalek, Petr [Joint Laboratory of Solid State Chemistry of Institute of Macromolecular Chemistry Academy of Sciences of Czech Republic, v.v.i., and University of Pardubice, University of Pardubice, Studentska 573, Pardubice, 532 10 (Czech Republic); Knotek, Petr [Department of General and Inorganic Chemistry, Faculty of Chemical Technology, University of Pardubice, Studentska 573, Pardubice, 532 10 Czech Republic (Czech Republic); Hromadko, Ludek; Macak, Jan M. [Center of Materials and Nanotechnologies, Faculty of Chemical Technology, University of Pardubice, Nam. Cs. Legii 565, Pardubice, 53002 (Czech Republic); Wagner, Tomas, E-mail: tomas.wagner@upce.cz [Department of General and Inorganic Chemistry, Faculty of Chemical Technology, University of Pardubice, Studentska 573, Pardubice, 532 10 Czech Republic (Czech Republic); Center of Materials and Nanotechnologies, Faculty of Chemical Technology, University of Pardubice, Nam. Cs. Legii 565, Pardubice, 53002 (Czech Republic)

    2016-09-30

    Highlights: • The resistive switching was studied from topological maps and spread current maps by conductive AFM. • Both surface particles and filaments were created under bias from conductive AFM. • The combination of topological map and spread current map proves the current did not flow through surface particles. • A model, consisting of interactions between charge carriers and Ag ions, were introduced to explain the experiment phenomena. - Abstract: In this paper, a study of resistive switching in Ag{sub x}AsS{sub 2} layer, based on a utilization of conductive atomic force microscope (AFM), is reported. As the result of biasing, two distinct regions were created on the surface (the conductive region and non-conductive region). Both were analysed from the spread current maps. The volume change, corresponding to the growth of Ag particles, was derived from the topological maps, recorded simultaneously with the current maps. Based on the results, a model explaining the mechanism of the Ag particle and Ag filament formation was proposed from the distribution of charge carriers and Ag ions.

  4. Large homogeneity ranges in the rare earth hydrides: a fiction to be revised

    International Nuclear Information System (INIS)

    Conder, K.; Longmei Wang; Boroch, E.; Kaldis, E.

    1991-01-01

    A large composition range of the solid solutions LnH 2 -LnH 3 (Ln=La, Ce) has been assumed for a long time. The structure of these solutions was believed to be cubic Fm3m with H atoms occupying tetrahedral and octahedral interstitials. Using x-ray diffraction and differential scanning calorimetry we have shown the existence of a large number of phases in both systems at T x ) are presented

  5. Fracture Growth Testing of Titanium 6AL-4V in AF-M315E

    Science.gov (United States)

    Sampson, Jeffrey W.; Martinez, Jonathan; McLean, Christopher

    2015-01-01

    The Green Propellant Infusion Mission (GPIM) will demonstrate the performance of AF-M315E monopropellant in orbit. Flight certification requires a safe-life analysis of the titanium alloy fuel tank to ensure inherent flaws will not cause failure during the design life. Material property inputs for this analysis require testing to determine the stress intensity factor for environmentally-assisted cracking (K (sub EAC)) of Ti 6Al-4V in combination with the AF-M315E monopropellant. Testing of single-edge notched specimens SE(B) representing the bulk tank membrane and weld material were performed in accordance with ASTM E1681. Specimens with fatigue pre-cracks were loaded into test fixtures so that the crack tips were exposed to the monopropellant at 50 degrees Centigrade for a duration of 1,000 hours. Specimens that did not fail during exposure were opened to inspect the crack surfaces for evidence of crack growth. The threshold stress intensity value, KEAC, is the highest applied stress intensity that produced neither a failure of the specimen during the exposure nor showed evidence of crack growth. The threshold stress intensity factor of the Ti 6Al-4V forged tank material when exposed to AF-M315E monopropellant was found to be at least 22.0 kilopounds per square inch. The stress intensity factor of the weld material was at least 31.3 kilopounds per square inch.

  6. In situ EC-AFM study of the effect of nanocrystals on the passivation and pit initiation in an Al-based metallic glass

    International Nuclear Information System (INIS)

    Zhang, S.D.; Liu, Z.W.; Wang, Z.M.; Wang, J.Q.

    2014-01-01

    Highlights: • The nanoscale corrosion on Al-rich glass was characterised by in situ EC-AFM. • The nanocrystals were identified from amorphous matrix by tapping mode AFM. • The formation of corrosion products is associated with the galvanic coupling. • The nanocrystals changed the local structure and component of the passive film. - Abstract: The effect of nanocrystals on pit initiation in metallic glasses is an unresolved issue. The passive film formation and pit initiation in the Al–Ni–Ce metallic glass were investigated using in situ electrochemical atomic force microscope (EC-AFM). The α-Al nanophases were identified from the amorphous matrix based upon the phase imaging in the tapping mode AFM. In the early stage of the passive film formation, the corrosion products Al(OH) 3 formed on the α-Al nanoparticles due to the galvanic coupling. The corrosion products incorporated into the passive film changed the local structure and component of the passive film, lowering its stability

  7. Miniature large range multi-axis force-torque sensor for biomechanical applications

    International Nuclear Information System (INIS)

    Brookhuis, R A; Sanders, R G P; Ma, K; Lammerink, T S J; De Boer, M J; Krijnen, G J M; Wiegerink, R J

    2015-01-01

    A miniature force sensor for the measurement of forces and moments at a human fingertip is designed and realized. Thin silicon pillars inside the sensor provide in-plane guidance for shear force measurement and provide the spring constant in normal direction. A corrugated silicon ring around the force sensitive area provides the spring constant in shear direction and seals the interior of the sensor. To detect all load components, capacitive read-out is used. A novel electrode pattern results in a large shear force sensitivity. The fingertip force sensor has a wide force range of up to 60 N in normal direction, ± 30 N in shear direction and a torque range of ± 25 N mm. (paper)

  8. Pathogen identification using peptide nanotube biosensors and impedance AFM

    Science.gov (United States)

    Maccuspie, Robert I.

    Pathogen identification at highly sensitive levels is crucial to meet urgent needs in fighting the spread of disease or detecting bioterrorism events. Toward that end, a new method for biosensing utilizing fluorescent antibody nanotubes is proposed. Fundamental studies on the self-assembly of these peptide nanotubes are performed, as are applications of aligning these nanotubes on surfaces. As biosensors, these nanotubes incorporate recognition units with antibodies at their ends and fluorescent signaling units at their sidewalls. When viral pathogens were mixed with these antibody nanotubes in solution, the nanotubes rapidly aggregated around the viruses. The size of the aggregates increased as the concentration of viruses increased, as detected by flow cytometry on the order of attomolar concentrations by changes in fluorescence and light scattering intensities. This enabled determination of the concentrations of viruses at trace levels (102 to 106 pfu/mL) within 30 minutes from the receipt of samples to the final quantitative data analysis, as demonstrated on Adenovirus, Herpes Simplex Virus, Influenza, and Vaccinia virus. As another separate approach, impedance AFM is used to study the electrical properties of individual viruses and nanoparticles used as model systems. The design, development, and implementation of the impedance AFM for an Asylum Research platform is described, as well as its application towards studying the impedance of individual nanoparticles as a model system for understanding the fundamental science of how the life cycle of a virus affects its electrical properties. In combination, these approaches fill a pressing need to quantify viruses both rapidly and sensitively.

  9. AFM assessment of the surface nano/microstructure on chemically damaged historical and model glasses

    International Nuclear Information System (INIS)

    Carmona, Noemi; Kowal, Andrzej; Rincon, Jesus-Maria; Villegas, Maria-Angeles

    2010-01-01

    Surface chemical damage on selected historical glasses from 13th to 19th centuries was evaluated by means of atomic force microscopy (AFM). Nano- and microstructure, roughness and topography of ancient glass samples have been compared with those of model glasses prepared by conventional melting at the laboratory with similar compositions to those most frequently found in historical glass pieces. The results obtained allow discussing the chemical degradation mechanisms in terms of the acid and/or basic chemical attack carried out by the combination of gaseous pollutants and environmental humidity. Even though deep corrosion features escape to the observation order of magnitude of the AF microscope used, the AFM technique proves to be quite useful for the study and evaluation of the most common surface pathologies of historical glasses with different compositions once submitted to natural weathering.

  10. AFM assessment of the surface nano/microstructure on chemically damaged historical and model glasses

    Energy Technology Data Exchange (ETDEWEB)

    Carmona, Noemi [Centro Nacional de Investigaciones Metalurgicas, CSIC, Avda. Gregorio del Amo, 8, 28040 Madrid (Spain); Kowal, Andrzej [Institute of Catalysis and Surface Chemistry, PAN, ul. Niezapominajek 8, 30239 Cracow (Poland); Rincon, Jesus-Maria [Instituto Eduardo Torroja de Ciencias de la Construccion, CSIC, C. Serrano Galvache s/n, 28033 Madrid (Spain); Villegas, Maria-Angeles, E-mail: mariangeles.villegas@cchs.csic.es [Centro Nacional de Investigaciones Metalurgicas, CSIC, Avda. Gregorio del Amo, 8, 28040 Madrid (Spain); Instituto de Historia, Centro de Ciencias Humanas y Sociales, CSIC, C. Albasanz, 26-28, 28037 Madrid (Spain)

    2010-01-15

    Surface chemical damage on selected historical glasses from 13th to 19th centuries was evaluated by means of atomic force microscopy (AFM). Nano- and microstructure, roughness and topography of ancient glass samples have been compared with those of model glasses prepared by conventional melting at the laboratory with similar compositions to those most frequently found in historical glass pieces. The results obtained allow discussing the chemical degradation mechanisms in terms of the acid and/or basic chemical attack carried out by the combination of gaseous pollutants and environmental humidity. Even though deep corrosion features escape to the observation order of magnitude of the AF microscope used, the AFM technique proves to be quite useful for the study and evaluation of the most common surface pathologies of historical glasses with different compositions once submitted to natural weathering.

  11. A Novel Atomic Force Microscope with Multi-Mode Scanner

    International Nuclear Information System (INIS)

    Qin, Chun; Zhang, Haijun; Xu, Rui; Han, Xu; Wang, Shuying

    2016-01-01

    A new type of atomic force microscope (AFM) with multi-mode scanner is proposed. The AFM system provides more than four scanning modes using a specially designed scanner with three tube piezoelectric ceramics and three stack piezoelectric ceramics. Sample scanning of small range with high resolution can be realized by using tube piezos, meanwhile, large range scanning can be achieved by stack piezos. Furthermore, the combination with tube piezos and stack piezos not only realizes high-resolution scanning of small samples with large- scale fluctuation structure, but also achieves small range area-selecting scanning. Corresponding experiments are carried out in terms of four different scanning modes showing that the AFM is of reliable stability, high resolution and can be widely applied in the fields of micro/nano-technology. (paper)

  12. Joint Research on Scatterometry and AFM Wafer Metrology

    Science.gov (United States)

    Bodermann, Bernd; Buhr, Egbert; Danzebrink, Hans-Ulrich; Bär, Markus; Scholze, Frank; Krumrey, Michael; Wurm, Matthias; Klapetek, Petr; Hansen, Poul-Erik; Korpelainen, Virpi; van Veghel, Marijn; Yacoot, Andrew; Siitonen, Samuli; El Gawhary, Omar; Burger, Sven; Saastamoinen, Toni

    2011-11-01

    Supported by the European Commission and EURAMET, a consortium of 10 participants from national metrology institutes, universities and companies has started a joint research project with the aim of overcoming current challenges in optical scatterometry for traceable linewidth metrology. Both experimental and modelling methods will be enhanced and different methods will be compared with each other and with specially adapted atomic force microscopy (AFM) and scanning electron microscopy (SEM) measurement systems in measurement comparisons. Additionally novel methods for sophisticated data analysis will be developed and investigated to reach significant reductions of the measurement uncertainties in critical dimension (CD) metrology. One final goal will be the realisation of a wafer based reference standard material for calibration of scatterometers.

  13. Force Spectroscopy of Hyaluronan by AFM; From H-bonded Networks Towards Single Chain Behavior

    NARCIS (Netherlands)

    Giannotti, M.I.; Rinaudo, Marguerite; Vancso, Gyula J.

    2007-01-01

    The conformational behavior of hyaluronan (HA) polysaccharide chains in aqueous NaCl solution was characterized directly at the single-molecule level. This comunication reports on one of the first single-chain atomic force microscopy (AFM) experiments performed at variable temperatures,

  14. AFM Imaging of Hybridization Chain Reaction Mediated Signal Transmission between Two DNA Origami Structures.

    Science.gov (United States)

    Helmig, Sarah; Gothelf, Kurt Vesterager

    2017-10-23

    Signal transfer is central to the controlled exchange of information in biology and advanced technologies. Therefore, the development of reliable, long-range signal transfer systems for artificial nanoscale assemblies is of great scientific interest. We have designed such a system for the signal transfer between two connected DNA nanostructures, using the hybridization chain reaction (HCR). Two sets of metastable DNA hairpins, one of which is immobilized at specific points along tracks on DNA origami structures, are polymerized to form a continuous DNA duplex, which is visible using atomic force microscopy (AFM). Upon addition of a designed initiator, the initiation signal is efficiently transferred more than 200 nm from a specific location on one origami structure to an end point on another origami structure. The system shows no significant loss of signal when crossing from one nanostructure to another and, therefore, has the potential to be applied to larger multi-component DNA assemblies. © 2017 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim.

  15. Capillary force on a tilted cylinder: Atomic Force Microscope (AFM) measurements.

    Science.gov (United States)

    Kosgodagan Acharige, Sébastien; Laurent, Justine; Steinberger, Audrey

    2017-11-01

    The capillary force in situations where the liquid meniscus is asymmetric, such as the one around a tilted object, has been hitherto barely investigated even though these situations are very common in practice. In particular, the capillary force exerted on a tilted object may depend on the dipping angle i. We investigate experimentally the capillary force that applies on a tilted cylinder as a function of its dipping angle i, using a home-built tilting Atomic Force Microscope (AFM) with custom made probes. A micrometric-size rod is glued at the end of an AFM cantilever of known stiffness, whose deflection is measured when the cylindrical probe is dipped in and retracted from reference liquids. We show that a torque correction is necessary to understand the measured deflection. We give the explicit expression of this correction as a function of the probes' geometrical parameters, so that its magnitude can be readily evaluated. The results are compatible with a vertical capillary force varying as 1/cosi, in agreement with a recent theoretical prediction. Finally, we discuss the accuracy of the method for measuring the surface tension times the cosine of the contact angle of the liquid on the probe. Copyright © 2017 Elsevier Inc. All rights reserved.

  16. Interaction and dynamics of ambient water adlayers on graphite probed using AFM voltage nanolithography and electrostatic force microscopy

    International Nuclear Information System (INIS)

    Gowthami, T; Raina, Gargi; Kurra, Narendra

    2014-01-01

    In this work, we report the impact of the interaction and dynamics of increasing ambient water adlayers on etch patterns on a hydrophobic highly oriented pyrolytic graphite (HOPG) surface obtained using atomic force microscopy (AFM) voltage nanolithography in contact mode by applying a positive bias to the sample. The changes in the dimensions of the etch patterns were investigated as a function of the increasing number of water adlayers present on the HOPG, which is varied by changing the time interval since HOPG cleavage. Changes in the width of the etch patterns and the surrounding water droplets were monitored with time, using intermittent-contact-mode AFM. Electrostatic force microscopy (EFM) has been employed to study the charged nature of the etch patterns and the neighboring water film with time. The width of the etch patterns made on freshly cleaved HOPG shows an increase of ∼33% over 48 h, whereas nine-day-old cleaved HOPG shows a 79% increase over the same period. No changes in the dimensions are observed while imaging in a nitrogen atmosphere soon after lithography. In ambient conditions, the EFM phase shift of the patterns shows a large change of ∼84–88% over 30 h. This study demonstrates the effect of the stored electrostatic energy of a polarized ice-like water adlayer, resulting in changes in the dimensions of the etch patterns long after lithography, whereas liquid-like water droplets do not affect the etch patterns. (paper)

  17. AFM characterization of protein net formation on a fibrous medium

    Directory of Open Access Journals (Sweden)

    Assis O.B.G.

    2000-01-01

    Full Text Available Lysozyme protein net is set on a glass fiber support using the self-assembly technique. Enzymatic film formation is followed by surface imaging via atomic force microscopy (AFM. Change in roughness as a function of deposition time is used as an indirect indicator of film formation. The objective was to form a protein film that would have no effect on the permeability of the medium, aiming at its application as a bioactive membrane or reactor suitable for bacteria and chemical interactions in aqueous media.

  18. Investigation of lateral forces in dynamic mode using combined AFM/STM

    OpenAIRE

    Atabak, Mehrdad

    2007-01-01

    Ankara : The Department of Physics and The Institute of Engineering and Science of Bilkent University, 2007. Thesis (Ph.D.) -- Bilkent University, 2007. Includes bibliographical references leaves 114-126. In this Ph.D. work, we constructed a ¯ber optic interferometer based non-contact Atomic Force Microscope (nc-AFM) combined with Scanning Tunneling Micro- scope(STM) to study lateral force interactions on Si(111)-(77) surface. The in- terferometer has been built in such a wa...

  19. Phase transition behaviors of the supported DPPC bilayer investigated by sum frequency generation (SFG) vibrational spectroscopy and atomic force microscopy (AFM).

    Science.gov (United States)

    Wu, Heng-Liang; Tong, Yujin; Peng, Qiling; Li, Na; Ye, Shen

    2016-01-21

    The phase transition behaviors of a supported bilayer of dipalmitoylphosphatidyl-choline (DPPC) have been systematically evaluated by in situ sum frequency generation (SFG) vibrational spectroscopy and atomic force microscopy (AFM). By using an asymmetric bilayer composed of per-deuterated and per-protonated monolayers, i.e., DPPC-d75/DPPC and a symmetric bilayer of DPPC/DPPC, we were able to probe the molecular structural changes during the phase transition process of the lipid bilayer by SFG spectroscopy. It was found that the DPPC bilayer is sequentially melted from the top (adjacent to the solution) to bottom leaflet (adjacent to the substrate) over a wide temperature range. The conformational ordering of the supported bilayer does not decrease (even slightly increases) during the phase transition process. The conformational defects in the bilayer can be removed after the complete melting process. The phase transition enthalpy for the bottom leaflet was found to be approximately three times greater than that for the top leaflet, indicating a strong interaction of the lipids with the substrate. The present SFG and AFM observations revealed similar temperature dependent profiles. Based on these results, the temperature-induced structural changes in the supported lipid bilayer during its phase transition process are discussed in comparison with previous studies.

  20. Effect of TMAH Etching Duration on the Formation of Silicon Nano wire Transistor Patterned by AFM Nano lithography

    International Nuclear Information System (INIS)

    Hutagalung, S.D.; Lew, K.C.

    2012-01-01

    Atomic force microscopy (AFM) lithography was applied to produce nano scale pattern for silicon nano wire transistor fabrication. This technique takes advantage of imaging facility of AFM and the ability of probe movement controlling over the sample surface to create nano patterns. A conductive AFM tip was used to grow the silicon oxide nano patterns on silicon on insulator (SOI) wafer. The applied tip-sample voltage and writing speed were well controlled in order to form pre-designed silicon oxide nano wire transistor structures. The effect of tetra methyl ammonium hydroxide (TMAH) etching duration on the oxide covered silicon nano wire transistor structure has been investigated. A completed silicon nano wire transistor was obtained by removing the oxide layer via hydrofluoric acid etching process. The fabricated silicon nano wire transistor consists of a silicon nano wire that acts as a channel with source and drain pads. A lateral gate pad with a nano wire head was fabricated very close to the channel in the formation of transistor structures. (author)

  1. Coating of AFM probes with aquatic humic and non-humic NOM to study their adhesion properties

    KAUST Repository

    Aubry, Cyril

    2013-06-01

    Atomic force microscopy (AFM) was used to study interaction forces between four Natural Organic Matter (NOM) samples of different physicochemical characteristics and origins and mica surface at a wide range of ionic strength. All NOM samples were strongly adsorbed on positively charged iron oxide-coated silica colloidal probe. Cross-sectioning by focused ion beam milling technique and elemental mapping by energy-filtered transmission electron microscopy indicated coating completeness of the NOM-coated colloidal probes. AFM-generated force-distance curves were analyzed to elucidate the nature and mechanisms of these interacting forces. Electrostatics and steric interactions were important contributors to repulsive forces during approach, although the latter became more influential with increasing ionic strength. Retracting force profiles showed a NOM adhesion behavior on mica consistent with its physicochemical characteristics. Humic-like substances, referred as the least hydrophilic NOM fraction, i.e., so called hydrophobic NOM, poorly adsorbed on hydrophilic mica due to their high content of ionized carboxyl groups and aromatic/hydrophobic character. However, adhesion force increased with increasing ionic strength, suggesting double layer compression. Conversely, polysaccharide-like substances showed high adhesion to mica. Hydrogen-bonding between hydroxyl groups on polysaccharide-like substances and highly electronegative elements on mica was suggested as the main adsorption mechanism, where the adhesion force decreased with increasing ionic strength. Results from this investigation indicated that all NOM samples retained their characteristics after the coating procedure. The experimental approach followed in this study can potentially be extended to investigate interactions between NOM and clean or fouled membranes as a function of NOM physicochemical characteristics and solution chemistry. © 2013 Elsevier Ltd.

  2. Polymorphism of amyloid fibrils formed by a peptide from the yeast prion protein Sup35: AFM and Tip-Enhanced Raman Scattering studies

    Energy Technology Data Exchange (ETDEWEB)

    Krasnoslobodtsev, Alexey V., E-mail: akrasnos@unomaha.edu [Department of Pharmaceutical Sciences, University of Nebraska Medical Center, 986025 Nebraska Medical Center, Omaha, NE 68198 (United States); Department of Physics, University of Nebraska Omaha, Omaha, NE 68182 (United States); Deckert-Gaudig, Tanja [IPHT-Leibniz Institute of Photonic Technology, Albert-Einstein-Str. 9, D-07745 Jena (Germany); Zhang, Yuliang [Department of Pharmaceutical Sciences, University of Nebraska Medical Center, 986025 Nebraska Medical Center, Omaha, NE 68198 (United States); Deckert, Volker [IPHT-Leibniz Institute of Photonic Technology, Albert-Einstein-Str. 9, D-07745 Jena (Germany); Institute for Physical Chemistry and Abbe Center of Photonics, University of Jena, Helmholtzweg 4, D-07743 Jena (Germany); Lyubchenko, Yuri L., E-mail: ylyubchenko@unmc.edu [Department of Pharmaceutical Sciences, University of Nebraska Medical Center, 986025 Nebraska Medical Center, Omaha, NE 68198 (United States)

    2016-06-15

    Aggregation of prion proteins is the cause of various prion related diseases. The infectious form of prions, amyloid aggregates, exist as multiple strains. The strains are thought to represent structurally different prion protein molecules packed into amyloid aggregates, but the knowledge on the structure of different types of aggregates is limited. Here we report on the use of AFM (Atomic Force Microscopy) and TERS (Tip-Enhanced Raman Scattering) to study morphological heterogeneity and access underlying conformational features of individual amyloid aggregates. Using AFM we identified the morphology of amyloid fibrils formed by the peptide (CGNNQQNY) from the yeast prion protein Sup35 that is critically involved in the aggregation of the full protein. TERS results demonstrate that morphologically different amyloid fibrils are composed of a distinct set of conformations. Fibrils formed at pH 5.6 are composed of a mixture of peptide conformations (β-sheets, random coil and α-helix) while fibrils formed in pH~2 solution primarily have β-sheets. Additionally, peak positions in the amide III region of the TERS spectra suggested that peptides have parallel arrangement of β-sheets for pH~2 fibrils and antiparallel arrangement for fibrils formed at pH 5.6. We also developed a methodology for detailed analysis of the peptide secondary structure by correlating intensity changes of Raman bands in different regions of TERS spectra. Such correlation established that structural composition of peptides is highly localized with large contribution of unordered secondary structures on a fibrillar surface. - Highlights: • Amyloid polymorphs were characterized by AFM and TERS. • A mixture of peptide secondary structures in fibrils were identified using TERS. • TERS recognizes packing arrangement (parallel versus antiparallel) of peptides. • TERS is a powerful tool for high resolution structural analysis of fibrils.

  3. AFM and uni-axial testing of pericardium exposed to radiotherapy doses

    Energy Technology Data Exchange (ETDEWEB)

    Daar, Eman, E-mail: e.daar@surrey.ac.uk [Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Kaabar, W. [Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Lei, C. [Division of Mechanical, Medical, and Aerospace Engineering, University of Surrey, Guildford GU2 7XH (United Kingdom); Keddie, J.L. [Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Nisbet, A. [Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Department of Medical Physics, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom); Bradley, D.A. [Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom)

    2011-10-01

    The pericardium, a double-layered sac that encloses the heart, is made up of collagen and elastin fibres embedded in an amorphous matrix (forming the extracellular matrix). Collagen fibres are aligned in multidirectional orientation layers. This free arrangement of fibres gives the pericardium its viscoelastic properties and the ability to deform in all directions. This is an important mechanical property for the heart to perform its physiological functions, acknowledging the fact that the heart is attached to different ligaments and muscles in all directions. The present study aims to investigate the effect of penetrating photon ionising radiation on bovine pericardium tissue. This links to an interest in seeking to understand possible mechanisms underlying cardiac complications following treatment of the left breast in radiotherapy regimes. Pericardium samples were subjected to doses in the range 0-80 Gy. Atomic force microscopy (AFM) has been applied in characterising changes in the infrastructural and mechanical properties of the tissues. Preliminary data for doses of 80 Gy shows there was no significant change in the D-spacing period of the banded structure collagen type I but a significant increase is observed in the FWHM of the fibril widths (by between 25% and 27%) over that of unirradiated pericardium tissue.

  4. Tribological behavior of micro/nano-patterned surfaces in contact with AFM colloidal probe

    International Nuclear Information System (INIS)

    Zhang Xiaoliang; Wang Xiu; Kong Wen; Yi Gewen; Jia Junhong

    2011-01-01

    In effort to investigate the influence of the micro/nano-patterning or surface texturing on the nanotribological properties of patterned surfaces, the patterned polydimethylsiloxane (PDMS) surfaces with pillars were fabricated by replica molding technique. The surface morphologies of patterned PDMS surfaces with varying pillar sizes and spacing between pillars were characterized by atomic force microscope (AFM) and scanning electron microscope (SEM). The AFM/FFM was used to acquire the friction force images of micro/nano-patterned surfaces using a colloidal probe. A difference in friction force produced a contrast on the friction force images when the colloidal probe slid over different regions of the patterned polymer surfaces. The average friction force of patterned surface was related to the spacing between the pillars and their size. It decreased with the decreasing of spacing between the pillars and the increasing of pillar size. A reduction in friction force was attributed to the reduced area of contact between patterned surface and colloidal probe. Additionally, the average friction force increased with increasing applied load and sliding velocity.

  5. Tribological behavior of micro/nano-patterned surfaces in contact with AFM colloidal probe

    Energy Technology Data Exchange (ETDEWEB)

    Zhang Xiaoliang; Wang Xiu; Kong Wen [State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000 (China); Graduate University of Chinese Academy of Sciences, Beijing 100049 (China); Yi Gewen [State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000 (China); Jia Junhong, E-mail: jhjia@licp.cas.cn [State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000 (China)

    2011-10-15

    In effort to investigate the influence of the micro/nano-patterning or surface texturing on the nanotribological properties of patterned surfaces, the patterned polydimethylsiloxane (PDMS) surfaces with pillars were fabricated by replica molding technique. The surface morphologies of patterned PDMS surfaces with varying pillar sizes and spacing between pillars were characterized by atomic force microscope (AFM) and scanning electron microscope (SEM). The AFM/FFM was used to acquire the friction force images of micro/nano-patterned surfaces using a colloidal probe. A difference in friction force produced a contrast on the friction force images when the colloidal probe slid over different regions of the patterned polymer surfaces. The average friction force of patterned surface was related to the spacing between the pillars and their size. It decreased with the decreasing of spacing between the pillars and the increasing of pillar size. A reduction in friction force was attributed to the reduced area of contact between patterned surface and colloidal probe. Additionally, the average friction force increased with increasing applied load and sliding velocity.

  6. Characterization of opto-electrical enhancement of tandem photoelectrochemical cells by using photoconductive-AFM

    Science.gov (United States)

    Park, Sun-Young; Elbersen, Rick; Huskens, Jurriaan; Gardeniers, Han; Lee, Joo-Yul; Mul, Guido; Heo, Jinhee

    2017-07-01

    Solar-to-hydrogen conversion by water splitting in photoelectrochemical cells (PECs) is a promising approach to alleviate problems associated with intermittency in solar energy supply and demand. Several interfacial resistances in photoelectrodes limit the performance of such cells, while the properties of interfaces are not easy to analyze in situ. We applied photoconductive-AFM to analyze the performance of WO3/p+n Si photoanodes, containing an ultra-thin metal interface of either Au or Pt. The Au interface consisted of Au nanoparticles with well-ordered interspacing, while Pt was present in the form of a continuous film. Photoconductive-AFM data show that upon illumination significantly larger currents are measured for the WO3/p+n Si anode equipped with the Au interface, as compared to the WO3/p+n Si anode with the Pt interface, in agreement with the better performance of the former electrode in a photoelectrochemical cell. The remarkable performance of the Au-containing electrode is proposed to be the result of favorable electron-hole recombination rates induced by the Au nanoparticles in a plasmon resonance excited state.

  7. A new approach of recognition of ellipsoidal micro- and nanoparticles on AFM images and determination of their sizes

    International Nuclear Information System (INIS)

    Akhmadeev, Albert A; Kh Salakhov, Myakzyum

    2016-01-01

    In this work we develop an approach of automatic recognition of ellipsoidal particles on the atomic force microscopy (AFM) image and determination of their size, which is based on image segmentation and the surface approximation by ellipsoids. In addition to the comparative simplicity and rapidity of processing, this method allows us to determine the size of particles, the surface of which is not completely visible on the image. The proposed method showed good results on simulated images including noisy ones. Using this algorithm the size distributions of silica particles on experimental AFM images have been determined. (paper)

  8. Thermally-treated Pt-coated silicon AFM tips for wear resistance in ferroelectric data storage

    International Nuclear Information System (INIS)

    Bhushan, Bharat; Palacio, Manuel; Kwak, Kwang Joo

    2008-01-01

    In ferroelectric data storage, a conductive atomic force microscopy (AFM) probe with a noble metal coating is placed in contact with a lead zirconate titanate (PZT) film. The understanding and improvement of probe tip wear, particularly at high velocities, is needed for high data rate recording. A commercial Pt-coated silicon AFM probe was thermally treated in order to form platinum silicide at the near-surface. Nanoindentation, nanoscratch and wear experiments were performed to evaluate the mechanical properties and wear performance at high velocities. The thermally treated tip exhibited lower wear than the untreated tip. The tip wear mechanism is adhesive and abrasive wear with some evidence of impact wear. The enhancement in mechanical properties and wear resistance in the thermally treated film is attributed to silicide formation in the near-surface. Auger electron spectroscopy and electrical resistivity measurements confirm the formation of platinum silicide. This study advances the understanding of thin film nanoscale surface interactions

  9. The use of functionalized AFM tips as molecular sensors in the detection of pesticides

    International Nuclear Information System (INIS)

    Deda, Daiana K.; Pereira, Barbara B.S.; Bueno, Carolina C.; Silva, Aline N. da; Ribeiro, Gabrielle A.; Amarante, Adriano M.; Leite, Fabio L.; Franca, Eduardo F.

    2013-01-01

    Atomic force spectroscopy, a technique derived from Atomic Force Microscopy (AFM), allowed to distinguish nonspecific and specific interactions between the acetolactate synthase enzyme (ALS) and anti-atrazine antibody biomolecules and the herbicides imazaquin, metsulfuron-methyl and atrazine. The presence of specific interactions increased the adhesion force (F adh ) between the AFM tip and the herbicides, which made the modified tip a powerful biosensor. Increases of approximately 132% and 145% in the F adh values were observed when a tip functionalized with ALS was used to detect imazaquin and metsulfuron-methyl, respectively. The presence of specific interactions between the atrazine and the anti-atrazine antibody also caused an increase in the F adh values (approximately 175%) compared to those observed when using an unfunctionalized tip. The molecular modeling results obtained with the ALS enzyme suggest that the orientation of the biomolecule on the tip surface could be suitable for allowing interaction with the herbicides imazaquin and metsulfuron-methyl. (author)

  10. The use of functionalized AFM tips as molecular sensors in the detection of pesticides

    Directory of Open Access Journals (Sweden)

    Daiana K. Deda

    2013-06-01

    Full Text Available Atomic force spectroscopy, a technique derived from Atomic Force Microscopy (AFM, allowed us to distinguish nonspecific and specific interactions between the acetolactate synthase enzyme (ALS and anti-atrazine antibody biomolecules and the herbicides imazaquin, metsulfuron-methyl and atrazine. The presence of specific interactions increased the adhesion force (Fadh between the AFM tip and the herbicides, which made the modified tip a powerful biosensor. Increases of approximately 132% and 145% in the Fadh values were observed when a tip functionalized with ALS was used to detect imazaquin and metsulfuron-methyl, respectively. The presence of specific interactions between the atrazine and the anti-atrazine antibody also caused an increase in the Fadh values (approximately 175% compared to those observed when using an unfunctionalized tip. The molecular modeling results obtained with the ALS enzyme suggest that the orientation of the biomolecule on the tip surface could be suitable for allowing interaction with the herbicides imazaquin and metsulfuron-methyl.

  11. The use of functionalized AFM tips as molecular sensors in the detection of pesticides

    Energy Technology Data Exchange (ETDEWEB)

    Deda, Daiana K.; Pereira, Barbara B.S.; Bueno, Carolina C.; Silva, Aline N. da; Ribeiro, Gabrielle A.; Amarante, Adriano M.; Leite, Fabio L., E-mail: fabioleite@ufscar.br [Universidade Federal de Sao Carlos (LNN/UFSCar), Sorocaba, SP (Brazil). Dept. de Fisica, Quimica e Matematica. Lab. de Nanoneurobiofisica; Franca, Eduardo F. [Universidade Federal de Uberlandia (UFU), MG (Brazil). Inst. de Quimica

    2013-11-01

    Atomic force spectroscopy, a technique derived from Atomic Force Microscopy (AFM), allowed to distinguish nonspecific and specific interactions between the acetolactate synthase enzyme (ALS) and anti-atrazine antibody biomolecules and the herbicides imazaquin, metsulfuron-methyl and atrazine. The presence of specific interactions increased the adhesion force (F{sub adh}) between the AFM tip and the herbicides, which made the modified tip a powerful biosensor. Increases of approximately 132% and 145% in the F{sub adh} values were observed when a tip functionalized with ALS was used to detect imazaquin and metsulfuron-methyl, respectively. The presence of specific interactions between the atrazine and the anti-atrazine antibody also caused an increase in the F{sub adh} values (approximately 175%) compared to those observed when using an unfunctionalized tip. The molecular modeling results obtained with the ALS enzyme suggest that the orientation of the biomolecule on the tip surface could be suitable for allowing interaction with the herbicides imazaquin and metsulfuron-methyl. (author)

  12. AFM studies of environmental effects on nanomechanical properties and cellular structure of human hair

    International Nuclear Information System (INIS)

    Bhushan, Bharat; Chen, Nianhuan

    2006-01-01

    Characterization of cellular structure and physical and mechanical properties of hair are essential to develop better cosmetic products and advance biological and cosmetic science. Although the morphology of the cellular structure of human hair has been traditionally investigated using scanning electron microscopy and transmission electron microscopy, these techniques provide limited capability to in situ study of the physical and mechanical properties of human hair in various environments. Atomic force microscopy (AFM) overcomes these problems and can be used for characterization in ambient conditions without requiring specific sample preparations and surface treatment. In this study, film thickness, adhesive forces and effective Young's modulus of various hair surfaces were measured at different environments (humidity and temperature) using force calibration plot technique with an AFM. Torsional resonance mode phase contrast images were also taken in order to characterize the morphology and cellular structure changes of human hair at different humidity. The correlation between the nanomechanical properties and the cellular structure of hair is discussed

  13. Local anodic oxidation by AFM tip developed for novel semiconductor nanodevices

    International Nuclear Information System (INIS)

    Cambel, Vladimir; Martaus, Jozef; Soltys, Jan; Kudela, Robert; Gregusova, Dagmar

    2008-01-01

    The local anodic oxidation (LAO) by the tip of atomic force microscope (AFM) is used for fabrication of nanometer-scaled structures and devices. We study the technology of LAO applied to semiconductor heterostructures, theoretically and experimentally as well. The goal is to improve the LAO process itself, i.e., to create narrow LAO lines that form high-energy barriers in the plane with the 2D electron gas. In the first part we show the electric field distribution in the system tip-sample during LAO. For samples with low-conductive cap layer the maximum electric field is shifted apart the tip apex, which leads to wide oxide lines. Our Monte Carlo (MC) calculations show how the height of the energy barrier in the system depends on the geometry of the created lines (trenches), and on voltage applied to the structure. Based on the calculations, we have proposed a novel LAO technology and applied it to InGaP/AlGaAs/GaAs heterostructure with doping layer only 6 nm beneath the surface. The doping layer can be oxidized easily by the AFM tip in this case, and the oxide objects can be removed by several etchants. This approach to the LAO technology leads to narrow LAO trenches (∼60 nm) and to energy barriers high enough for room- and low-temperature applications

  14. AFM of the ultrastructural and mechanical properties of lipid-raft-disrupted and/or cold-treated endothelial cells.

    Science.gov (United States)

    Wu, Li; Huang, Jie; Yu, Xiaoxue; Zhou, Xiaoqing; Gan, Chaoye; Li, Ming; Chen, Yong

    2014-02-01

    The nonionic detergent extraction at 4 °C and the cholesterol-depletion-induced lipid raft disruption are the two widely used experimental strategies for lipid raft research. However, the effects of raft disruption and/or cold treatment on the ultrastructural and mechanical properties of cells are still unclear. Here, we evaluated the effects of raft disruption and/or cold (4 °C) treatment on these properties of living human umbilical vein endothelial cells (HUVECs). At first, the cholesterol-depletion-induced raft disruption was visualized by confocal microscopy and atomic force microscopy (AFM) in combination with fluorescent quantum dots. Next, the cold-induced cell contraction and the formation of end-branched filopodia were observed by confocal microscopy and AFM. Then, the cell-surface ultrastructures were imaged by AFM, and the data showed that raft disruption and cold treatment induced opposite effects on cell-surface roughness (a significant decrease and a significant increase, respectively). Moreover, the cell-surface mechanical properties (stiffness and adhesion force) of raft-disrupted- and/or cold-treated HUVECs were measured by the force measurement function of AFM. We found that raft disruption and cold treatment induced parallel effects on cell stiffness (increase) or adhesion force (decrease) and that the combination of the two treatments caused dramatically strengthened effects. Finally, raft disruption was found to significantly impair cell migration as previously reported, whereas temporary cold treatment only caused a slight but nonsignificant decrease in cell migration performed at physiological temperature. Although the mechanisms for causing these results might be complicated and more in-depth studies will be needed, our data may provide important information for better understanding the effects of raft disruption or cold treatment on cells and the two strategies for lipid raft research.

  15. AFM and SEM study of the effects of etching on IPS-Empress 2 TM dental ceramic

    Science.gov (United States)

    Luo, X.-P.; Silikas, N.; Allaf, M.; Wilson, N. H. F.; Watts, D. C.

    2001-10-01

    The aim of this study was to investigate the effects of increasing etching time on the surface of the new dental material, IPS-Empress 2 TM glass ceramic. Twenty one IPS-Empress 2 TM glass ceramic samples were made from IPS-Empress 2 TM ingots through lost-wax, hot-pressed ceramic fabrication technology. All samples were highly polished and cleaned ultrasonically for 5 min in acetone before and after etching with 9.6% hydrofluoric acid gel. The etching times were 0, 10, 20, 30, 60, 90 and 120 s respectively. Microstructure was analysed by scanning electron microscopy (SEM) and atomic force microscopy (AFM) was used to evaluate the surface roughness and topography. Observations with SEM showed that etching with hydrofluoric acid resulted in preferential dissolution of glass matrix, and that partially supported crystals within the glass matrix were lost with increasing etching time. AFM measurements indicated that etching increased the surface roughness of the glass-ceramic. A simple least-squares linear regression was used to establish a relationship between surface roughness parameters ( Ra, RMS), and etching time, for which r2>0.94. This study demonstrates the benefits of combining two microscopic methods for a better understanding of the surface. SEM showed the mode of action of hydrofluoric acid on the ceramic and AFM provided valuable data regarding the extent of surface degradation relative to etching time.

  16. Coating of AFM probes with aquatic humic and non-humic NOM to study their adhesion properties

    KAUST Repository

    Aubry, Cyril; Gutié rrez, Leonardo A.; Croue, Jean-Philippe

    2013-01-01

    -coated colloidal probes. AFM-generated force-distance curves were analyzed to elucidate the nature and mechanisms of these interacting forces. Electrostatics and steric interactions were important contributors to repulsive forces during approach, although

  17. Investigation of the influence of UV irradiation on collagen thin films by AFM imaging

    Energy Technology Data Exchange (ETDEWEB)

    Stylianou, Andreas, E-mail: styliand@mail.ntua.gr; Yova, Dido; Alexandratou, Eleni

    2014-12-01

    Collagen is the major fibrous extracellular matrix protein and due to its unique properties, it has been widely used as biomaterial, scaffold and cell-substrate. The aim of the paper was to use Atomic Force Microscopy (AFM) in order to investigate well-characterized collagen thin films after ultraviolet light (UV) irradiation. The films were also used as in vitro culturing substrates in order to investigate the UV-induced alterations to fibroblasts. A special attention was given in the alteration on collagen D-periodicity. For short irradiation times, spectroscopy (fluorescence/absorption) studies demonstrated that photodegradation took place and AFM imaging showed alterations in surface roughness. Also, it was highlighted that UV-irradiation had different effects when it was applied on collagen solution than on films. Concerning fibroblast culturing, it was shown that fibroblast behavior was affected after UV irradiation of both collagen solution and films. Furthermore, after a long irradiation time, collagen fibrils were deformed revealing that collagen fibrils are consisting of multiple shells and D-periodicity occurred on both outer and inner shells. The clarification of the effects of UV light on collagen and the induced modifications of cell behavior on UV-irradiated collagen-based surfaces will contribute to the better understanding of cell–matrix interactions in the nanoscale and will assist in the appropriate use of UV light for sterilizing and photo-cross-linking applications. - Highlights: • Collagen thin films were formed and exposed in UV irradiation. • Collagen thin films were formed from UV-irradiated collagen solution. • Nanocharacterization of collagen thin films by AFM • Fluorescence and absorption spectroscopy studies on collagen films • Investigation of fibroblast response on collagen films.

  18. Investigation of the influence of UV irradiation on collagen thin films by AFM imaging

    International Nuclear Information System (INIS)

    Stylianou, Andreas; Yova, Dido; Alexandratou, Eleni

    2014-01-01

    Collagen is the major fibrous extracellular matrix protein and due to its unique properties, it has been widely used as biomaterial, scaffold and cell-substrate. The aim of the paper was to use Atomic Force Microscopy (AFM) in order to investigate well-characterized collagen thin films after ultraviolet light (UV) irradiation. The films were also used as in vitro culturing substrates in order to investigate the UV-induced alterations to fibroblasts. A special attention was given in the alteration on collagen D-periodicity. For short irradiation times, spectroscopy (fluorescence/absorption) studies demonstrated that photodegradation took place and AFM imaging showed alterations in surface roughness. Also, it was highlighted that UV-irradiation had different effects when it was applied on collagen solution than on films. Concerning fibroblast culturing, it was shown that fibroblast behavior was affected after UV irradiation of both collagen solution and films. Furthermore, after a long irradiation time, collagen fibrils were deformed revealing that collagen fibrils are consisting of multiple shells and D-periodicity occurred on both outer and inner shells. The clarification of the effects of UV light on collagen and the induced modifications of cell behavior on UV-irradiated collagen-based surfaces will contribute to the better understanding of cell–matrix interactions in the nanoscale and will assist in the appropriate use of UV light for sterilizing and photo-cross-linking applications. - Highlights: • Collagen thin films were formed and exposed in UV irradiation. • Collagen thin films were formed from UV-irradiated collagen solution. • Nanocharacterization of collagen thin films by AFM • Fluorescence and absorption spectroscopy studies on collagen films • Investigation of fibroblast response on collagen films

  19. Tumor suppressor protein SMAR1 modulates the roughness of cell surface: combined AFM and SEM study

    Directory of Open Access Journals (Sweden)

    Mamgain Hitesh

    2009-10-01

    Full Text Available Abstract Background Imaging tools such as scanning electron microscope (SEM and atomic force microscope (AFM can be used to produce high-resolution topographic images of biomedical specimens and hence are well suited for imaging alterations in cell morphology. We have studied the correlation of SMAR1 expression with cell surface smoothness in cell lines as well as in different grades of human breast cancer and mouse tumor sections. Methods We validated knockdown and overexpression of SMAR1 using RT-PCR as well as Western blotting in human embryonic kidney (HEK 293, human breast cancer (MCF-7 and mouse melanoma (B16F1 cell lines. The samples were then processed for cell surface roughness studies using atomic force microscopy (AFM and scanning electron microscopy (SEM. The same samples were used for microarray analysis as well. Tumors sections from control and SMAR1 treated mice as well as tissues sections from different grades of human breast cancer on poly L-lysine coated slides were used for AFM and SEM studies. Results Tumor sections from mice injected with melanoma cells showed pronounced surface roughness. In contrast, tumor sections obtained from nude mice that were first injected with melanoma cells followed by repeated injections of SMAR1-P44 peptide, exhibited relatively smoother surface profile. Interestingly, human breast cancer tissue sections that showed reduced SMAR1 expression exhibited increased surface roughness compared to the adjacent normal breast tissue. Our AFM data establishes that treatment of cells with SMAR1-P44 results into increase in cytoskeletal volume that is supported by comparative gene expression data showing an increase in the expression of specific cytoskeletal proteins compared to the control cells. Altogether, these findings indicate that tumor suppressor function of SMAR1 might be exhibited through smoothening of cell surface by regulating expression of cell surface proteins. Conclusion Tumor suppressor

  20. AFM Colloidal Probe Measurements Implicate Capillary Condensation in Punch-Particle Surface Interactions during Tableting.

    Science.gov (United States)

    Badal Tejedor, Maria; Nordgren, Niklas; Schuleit, Michael; Millqvist-Fureby, Anna; Rutland, Mark W

    2017-11-21

    Adhesion of the powders to the punches is a common issue during tableting. This phenomenon is known as sticking and affects the quality of the manufactured tablets. Defective tablets increase the cost of the manufacturing process. Thus, the ability to predict the tableting performance of the formulation blend before the process is scaled-up is important. The adhesive propensity of the powder to the tableting tools is mostly governed by the surface-surface adhesive interactions. Atomic force microscopy (AFM) colloidal probe is a surface characterization technique that allows the measurement of the adhesive interactions between two materials of interest. In this study, AFM steel colloidal probe measurements were performed on ibuprofen, MCC (microcrystalline cellulose), α-lactose monohydrate, and spray-dried lactose particles as an approach to modeling the punch-particle surface interactions during tableting. The excipients (lactose and MCC) showed constant, small, attractive, and adhesive forces toward the steel surface after a repeated number of contacts. In comparison, ibuprofen displayed a much larger attractive and adhesive interaction increasing over time both in magnitude and in jump-in/jump-out separation distance. The type of interaction acting on the excipient-steel interface can be related to a van der Waals force, which is relatively weak and short-ranged. By contrast, the ibuprofen-steel interaction is described by a capillary force profile. Even though ibuprofen is not highly hydrophilic, the relatively smooth surfaces of the crystals allow "contact flooding" upon contact with the steel probe. Capillary forces increase because of the "harvesting" of moisture-due to the fast condensation kinetics-leaving a residual condensate that contributes to increase the interaction force after each consecutive contact. Local asperity contacts on the more hydrophilic surface of the excipients prevent the flooding of the contact zone, and there is no such adhesive

  1. Radiolysis of aqueous solutions of ammonium bicarbonate over a large dose range

    International Nuclear Information System (INIS)

    Draganic, Z.D.; Negron-Mendoza, A.; Vujosevic, S.I.; Navarro-Gonzales, R.; Albarran-Sanchez, M.G.

    1991-01-01

    Oxygen-free aqueous solutions of 0.05 mol dm -3 ammonium and sodium bicarbonate were studied after receiving various doses of 60 Co gammas (0.01-400 kGy) or 0.5-20 Gy pulses of 10 Mev electrons. Formate and oxalate were found to be the main radiolytic products, in addition to trace amounts of formaldehyde and an unidentified polymer. A large initial yield of formate in the γ-radiolysis, G(HCOO - ) = 2.2, is due to the reaction COO - + HCO 3 - ↔ HCOO - +CO 3 - . The efficiency of organic synthesis within the large dose range studied is low and is explained by efficient pathways leading to the reformation of bicarbonate, where the reaction COO - + CO 3 - is particularly significant. (author)

  2. Friction behavior of nano-textured polyimide surfaces measured by AFM colloidal probe

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, Xiaoliang [College of Equipment Manufacturing, Hebei University of Engineering, Handan 056038 (China); State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000 (China); Wu, Chunxia; Che, Hongwei; Hou, Junxian [College of Equipment Manufacturing, Hebei University of Engineering, Handan 056038 (China); Jia, Junhong, E-mail: jhjia@licp.cas.cn [State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000 (China)

    2014-11-30

    Highlights: • Flat PI film and nano-textured PI film were prepared by spin-coating process. • The nano-textured PI surface has effectively reduced the adhesion and friction. • Friction increased with the increasing of contact area and adhesion. • The growth rate of friction decreased with the increasing of applied load. - Abstract: Flat polyimide (PI) film and silicon dioxide nanoparticle-textured PI film were prepared by means of the spin-coating technique. The adhesion and friction properties of the flat PI surface and nano-textured PI surface were investigated by a series of Atomic force microscope (AFM) colloidal probes. Experimental results revealed that the nano-textured PI surface can significantly reduce the adhesive force and friction force, compared with the flat PI surface. The main reason is that the nano-textures can reduce the contact area between the sample surface and colloidal probe. The effect of colloidal probe size on the friction behavior of the flat and nano-textured PI surfaces was evaluated. The adhesive force and friction force of nano-textured PI surface were increased with the increasing of the size of interacting pairs (AFM colloidal probe) due to the increased contact area. Moreover, the friction forces of flat and nano-textured PI surfaces were increased with applied load and sliding velocity.

  3. Adhesive interaction measured between AFM probe and lung epithelial type II cells

    International Nuclear Information System (INIS)

    Leonenko, Zoya; Finot, Eric; Amrein, Matthias

    2007-01-01

    The toxicity of inhaled nanoparticles entering the body through the lung is thought to be initially defined by the electrostatic and adhesive interaction of the particles with lung's wall. Here, we investigated the first step of the interaction of nanoparticles with lung epithelial cells using atomic force microscope (AFM) as a force apparatus. Nanoparticles were modeled by the apex of the AFM tip and the forces of interaction between the tip and the cell analyzed over time. The adhesive force and work of adhesion strongly increased for the first 100 s of contact and then leveled out. During this time, the tip was penetrating deeply into the cell. It first crossed a stiff region of the cell and then entered a much more compliant cell region. The work of adhesion and its progression over time were not dependent on the load with which the tip was brought into contact with the cell. We conclude that the initial thermodynamic aspects and the time course of the uptake of nanoparticles by lung epithelial cells can be studied using our experimental approach. It is discussed how the potential health threat posed by nanoparticles of different size and surface characteristics can be evaluated using the method presented

  4. Macro-SICM: A Scanning Ion Conductance Microscope for Large-Range Imaging.

    Science.gov (United States)

    Schierbaum, Nicolas; Hack, Martin; Betz, Oliver; Schäffer, Tilman E

    2018-04-17

    The scanning ion conductance microscope (SICM) is a versatile, high-resolution imaging technique that uses an electrolyte-filled nanopipet as a probe. Its noncontact imaging principle makes the SICM uniquely suited for the investigation of soft and delicate surface structures in a liquid environment. The SICM has found an ever-increasing number of applications in chemistry, physics, and biology. However, a drawback of conventional SICMs is their relatively small scan range (typically 100 μm × 100 μm in the lateral and 10 μm in the vertical direction). We have developed a Macro-SICM with an exceedingly large scan range of 25 mm × 25 mm in the lateral and 0.25 mm in the vertical direction. We demonstrate the high versatility of the Macro-SICM by imaging at different length scales: from centimeters (fingerprint, coin) to millimeters (bovine tongue tissue, insect wing) to micrometers (cellular extensions). We applied the Macro-SICM to the study of collective cell migration in epithelial wound healing.

  5. Magnetoelectric versus thermal actuation characteristics of shear force AFM probes with piezoresistive detection

    International Nuclear Information System (INIS)

    Sierakowski, Andrzej; Janus, Paweł; Dobrowolski, Rafał; Grabiec, Piotr; Kopiec, Daniel; Majstrzyk, Wojciech; Kunicki, Piotr; Gotszalk, Teodor; Rangelow, Ivo W

    2017-01-01

    In this paper the authors compare methods used for piezoresistive microcantilevers actuation for the atomic force microscopy (AFM) imaging in the dynamic shear force mode. The piezoresistive detection is an attractive technique comparing the optical beam detection of deflection. The principal advantage is that no external alignment of optical source and detector are needed. When the microcantilever is deflected, the stress is transferred into a change of resistivity of piezoresistors. The integration of piezoresistive read-out provides a promising solution in realizing a compact non-contact AFM. Resolution of piezoresistive read-out is limited by three main noise sources: Johnson, 1/ f and thermomechanical noise. In the dynamic shear force mode measurement the method used for cantilever actuation will also affect the recorded noise in the piezoresistive detection circuit. This is the result of a crosstalk between an aluminium path (current loop used for actuation) and piezoresistors located near the base of the beam. In this paper authors described an elaborated in ITE (Institute of Electron Technology) technology of fabrication cantilevers with piezoresistive detection of deflection and compared efficiency of two methods used for cantilever actuation. (paper)

  6. Vickers Hardness of Diamond and cBN Single Crystals: AFM Approach

    Directory of Open Access Journals (Sweden)

    Sergey Dub

    2017-12-01

    Full Text Available Atomic force microscopy in different operation modes (topography, derivative topography, and phase contrast was used to obtain 3D images of Vickers indents on the surface of diamond and cBN single crystals with high spatial resolution. Confocal Raman spectroscopy and Kelvin probe force microscopy were used to study the structure of the material in the indents. It was found that Vickers indents in diamond has no sharp and clear borders. However, the phase contrast operation mode of the AFM reveals a new viscoelastic phase in the indent in diamond. Raman spectroscopy and Kelvin probe force microscopy revealed that the new phase in the indent is disordered graphite, which was formed due to the pressure-induced phase transformation in the diamond during the hardness test. The projected contact area of the graphite layer in the indent allows us to measure the Vickers hardness of type-Ib synthetic diamond. In contrast to diamond, very high plasticity was observed for 0.5 N load indents on the (001 cBN single crystal face. Radial and ring cracks were absent, the shape of the indents was close to a square, and there were linear details in the indent, which looked like slip lines. The Vickers hardness of the (111 synthetic diamond and (111 and (001 cBN single crystals were determined using the AFM images and with account for the elastic deformation of the diamond Vickers indenter during the tests.

  7. A compact, large-range interferometer for precision measurement and inertial sensing

    Science.gov (United States)

    Cooper, S. J.; Collins, C. J.; Green, A. C.; Hoyland, D.; Speake, C. C.; Freise, A.; Mow-Lowry, C. M.

    2018-05-01

    We present a compact, fibre-coupled interferometer with high sensitivity and a large working range. We propose to use this interferometer as a readout mechanism for future inertial sensors, removing a major limiting noise source, and in precision positioning systems. The interferometer’s peak sensitivity is 2 × 10-{14} m \\sqrt{Hz-1} at 70 Hz and 7 × 10-{11} m \\sqrt{Hz-1} at 10 mHz. If deployed on a GS-13 geophone, the resulting inertial sensing output will be limited by the suspension thermal noise of the reference mass from 10 mHz to 2 Hz.

  8. Nanobiosensors Based on Chemically Modified AFM Probes: A Useful Tool for Metsulfuron-Methyl Detection

    Directory of Open Access Journals (Sweden)

    Fabio L. Leite

    2013-01-01

    Full Text Available The use of agrochemicals has increased considerably in recent years, and consequently, there has been increased exposure of ecosystems and human populations to these highly toxic compounds. The study and development of methodologies to detect these substances with greater sensitivity has become extremely relevant. This article describes, for the first time, the use of atomic force spectroscopy (AFS in the detection of enzyme-inhibiting herbicides. A nanobiosensor based on an atomic force microscopy (AFM tip functionalised with the acetolactate synthase (ALS enzyme was developed and characterised. The herbicide metsulfuron-methyl, an ALS inhibitor, was successfully detected through the acquisition of force curves using this biosensor. The adhesion force values were considerably higher when the biosensor was used. An increase of ~250% was achieved relative to the adhesion force using an unfunctionalised AFM tip. This considerable increase was the result of a specific interaction between the enzyme and the herbicide, which was primarily responsible for the efficiency of the nanobiosensor. These results indicate that this methodology is promising for the detection of herbicides, pesticides, and other environmental contaminants.

  9. Characterization of Bitumen Micro-Mechanical Behaviors Using AFM, Phase Dynamics Theory and MD Simulation

    Directory of Open Access Journals (Sweden)

    Yue Hou

    2017-02-01

    Full Text Available Fundamental understanding of micro-mechanical behaviors in bitumen, including phase separation, micro-friction, micro-abrasion, etc., can help the pavement engineers better understand the bitumen mechanical performances at macroscale. Recent researches show that the microstructure evolution in bitumen will directly affect its surface structure and micro-mechanical performance. In this study, the bitumen microstructure and micro-mechanical behaviors are studied using Atomic Force Microscopy (AFM experiments, Phase Dynamics Theory and Molecular Dynamics (MD Simulation. The AFM experiment results show that different phase-structure will occur at the surface of the bitumen samples under certain thermodynamic conditions at microscale. The phenomenon can be explained using the phase dynamics theory, where the effects of stability parameter and temperature on bitumen microstructure and micro-mechanical behavior are studied combined with MD Simulation. Simulation results show that the saturates phase, in contrast to the naphthene aromatics phase, plays a major role in bitumen micro-mechanical behavior. A high stress zone occurs at the interface between the saturates phase and the naphthene aromatics phase, which may form discontinuities that further affect the bitumen frictional performance.

  10. Characterization of Bitumen Micro-Mechanical Behaviors Using AFM, Phase Dynamics Theory and MD Simulation.

    Science.gov (United States)

    Hou, Yue; Wang, Linbing; Wang, Dawei; Guo, Meng; Liu, Pengfei; Yu, Jianxin

    2017-02-21

    Fundamental understanding of micro-mechanical behaviors in bitumen, including phase separation, micro-friction, micro-abrasion, etc., can help the pavement engineers better understand the bitumen mechanical performances at macroscale. Recent researches show that the microstructure evolution in bitumen will directly affect its surface structure and micro-mechanical performance. In this study, the bitumen microstructure and micro-mechanical behaviors are studied using Atomic Force Microscopy (AFM) experiments, Phase Dynamics Theory and Molecular Dynamics (MD) Simulation. The AFM experiment results show that different phase-structure will occur at the surface of the bitumen samples under certain thermodynamic conditions at microscale. The phenomenon can be explained using the phase dynamics theory, where the effects of stability parameter and temperature on bitumen microstructure and micro-mechanical behavior are studied combined with MD Simulation. Simulation results show that the saturates phase, in contrast to the naphthene aromatics phase, plays a major role in bitumen micro-mechanical behavior. A high stress zone occurs at the interface between the saturates phase and the naphthene aromatics phase, which may form discontinuities that further affect the bitumen frictional performance.

  11. Simulated structure and imaging of NTCDI on Si(1 1 1)-7 × 7 : a combined STM, NC-AFM and DFT study

    Science.gov (United States)

    Jarvis, S. P.; Sweetman, A. M.; Lekkas, I.; Champness, N. R.; Kantorovich, L.; Moriarty, P.

    2015-02-01

    The adsorption of naphthalene tetracarboxylic diimide (NTCDI) on Si(1 1 1)-7 × 7 is investigated through a combination of scanning tunnelling microscopy (STM), noncontact atomic force microscopy (NC-AFM) and density functional theory (DFT) calculations. We show that NTCDI adopts multiple planar adsorption geometries on the Si(1 1 1)-7 × 7 surface which can be imaged with intramolecular bond resolution using NC-AFM. DFT calculations reveal adsorption is dominated by covalent bond formation between the molecular oxygen atoms and the surface silicon adatoms. The chemisorption of the molecule is found to induce subtle distortions to the molecular structure, which are observed in NC-AFM images.

  12. Dual frequency modulation with two cantilevers in series: a possible means to rapidly acquire tip–sample interaction force curves with dynamic AFM

    International Nuclear Information System (INIS)

    Solares, Santiago D; Chawla, Gaurav

    2008-01-01

    One common application of atomic force microscopy (AFM) is the acquisition of tip–sample interaction force curves. However, this can be a slow process when the user is interested in studying non-uniform samples, because existing contact- and dynamic-mode methods require that the measurement be performed at one fixed surface point at a time. This paper proposes an AFM method based on dual frequency modulation using two cantilevers in series, which could be used to measure the tip–sample interaction force curves and topography of the entire sample with a single surface scan, in a time that is comparable to the time needed to collect a topographic image with current AFM imaging modes. Numerical simulation results are provided along with recommended parameters to characterize tip–sample interactions resembling those of conventional silicon tips and carbon nanotube tips tapping on silicon surfaces

  13. Similar range of motion and function after resurfacing large-head or standard total hip arthroplasty

    DEFF Research Database (Denmark)

    Penny, Jeannette Østergaard; Ovesen, Ole; Varmarken, Jens-Erik

    2013-01-01

    BACKGROUND AND PURPOSE: Large-size hip articulations may improve range of motion (ROM) and function compared to a 28-mm THA, and the low risk of dislocation allows the patients more activity postoperatively. On the other hand, the greater extent of surgery for resurfacing hip arthroplasty (RHA......° (35), 232° (36), and 225° (30) respectively, but the differences were not statistically significant. The 3 groups were similar regarding Harris hip score, UCLA activity score, step rate, and sick leave. INTERPRETATION: Head size had no influence on range of motion. The lack of restriction allowed...... for large articulations did not improve the clinical and patient-perceived outcomes. The more extensive surgical procedure of RHA did not impair the rehabilitation. This project is registered at ClinicalTrials.gov under # NCT01113762....

  14. STM, SECPM, AFM and Electrochemistry on Single Crystalline Surfaces

    Directory of Open Access Journals (Sweden)

    Ulrich Stimming

    2010-08-01

    Full Text Available Scanning probe microscopy (SPM techniques have had a great impact on research fields of surface science and nanotechnology during the last decades. They are used to investigate surfaces with scanning ranges between several 100 mm down to atomic resolution. Depending on experimental conditions, and the interaction forces between probe and sample, different SPM techniques allow mapping of different surface properties. In this work, scanning tunneling microscopy (STM in air and under electrochemical conditions (EC-STM, atomic force microscopy (AFM in air and scanning electrochemical potential microscopy (SECPM under electrochemical conditions, were used to study different single crystalline surfaces in electrochemistry. Especially SECPM offers potentially new insights into the solid-liquid interface by providing the possibility to image the potential distribution of the surface, with a resolution that is comparable to STM. In electrocatalysis, nanostructured catalysts supported on different electrode materials often show behavior different from their bulk electrodes. This was experimentally and theoretically shown for several combinations and recently on Pt on Au(111 towards fuel cell relevant reactions. For these investigations single crystals often provide accurate and well defined reference and support systems. We will show heteroepitaxially grown Ru, Ir and Rh single crystalline surface films and bulk Au single crystals with different orientations under electrochemical conditions. Image studies from all three different SPM methods will be presented and compared to electrochemical data obtained by cyclic voltammetry in acidic media. The quality of the single crystalline supports will be verified by the SPM images and the cyclic voltammograms. Furthermore, an outlook will be presented on how such supports can be used in electrocatalytic studies.

  15. Fully low voltage and large area searching scanning tunneling microscope

    International Nuclear Information System (INIS)

    Pang, Zongqiang; Wang, Jihui; Lu, Qingyou

    2009-01-01

    We present a novel scanning tunneling microscope (STM), which allows the tip to travel a large distance (millimeters) on the sample and take images (to find microscopic targets) anywhere it reaches without losing atomic resolution. This broad range searching capability, together with the coarse approach and scan motion, is all done with only one single piezoelectric tube scanner as well as with only low voltages (<15 V). Simple structure, low interference and high precision are thus achieved. To this end, a pillar and a tube scanner are mounted in parallel on a base with one ball glued on the pillar top and two balls glued on the scanner top. These three balls form a narrow triangle, which supports a triangular slider piece. By inertial stepping, the scanner can move the slider toward the pillar (coarse approach) or rotate the slider about the pillar (travel along sample surface). Since all the stepping motions are driven by the scanner's lateral bending which is large per unit voltage, high voltages are unnecessary. The technology is also applicable to scanning force microscopes (SFM) such as atomic force microscopes (AFM), etc

  16. Evaluation of elastic properties of DLC layers using resonant ultrasound spectroscopy and AFM nanoindentation

    Czech Academy of Sciences Publication Activity Database

    Kocourek, Tomáš; Růžek, Michal; Landa, Michal; Jelínek, Miroslav; Mikšovský, Jan; Kopeček, Jaromír

    2011-01-01

    Roč. 205, č. 2 (2011), S67-S70 ISSN 0257-8972 R&D Projects: GA ČR(CZ) GA101/09/0702 Institutional research plan: CEZ:AV0Z10100522; CEZ:AV0Z20760514 Keywords : RUS-resonant ultrasound spectroscopy * PLD * diamond-like carbon * elastic properties * AFM nanoindentation Subject RIV: BH - Optics, Masers, Lasers Impact factor: 1.867, year: 2011

  17. Iron in carbonate containing AFm phases

    International Nuclear Information System (INIS)

    Dilnesa, B.Z.; Lothenbach, B.; Le Saout, G.; Renaudin, G.; Mesbah, A.; Filinchuk, Y.; Wichser, A.; Wieland, E.

    2011-01-01

    One of the AFm phases in hydrated Portland cement is Ca 3 (Al x Fe 2 - x )O 6 .CaCO 3 .nH 2 O. It is based on hexagonal and platey structural elements and the interlayer structure incorporates CO 3 2- . The solid phases were experimentally synthesized and characterized by different techniques including X-ray techniques (XRD and EXAFS) and vibrational spectroscopy techniques (IR, Raman). Fe-monocarbonate (Fe-Mc) and Al-monocarbonate (Al-Mc) were found to be stable up to 50 o C, while Fe-hemicarbonate (Fe-Hc) was unstable with respect to Fe-Mc in the presence of calcite. Fe-Mc has a rhombohedral R3-barc symmetry which is different from the triclinic of the Al analogue. Both XRD and thermodynamic modelling of the liquid compositions indicated that Al-Mc and the Fe-Mc phases do not form solid solution. The solubility products were calculated experimentally at 20 o C and 50 o C. Under standards condition the solubility products and other thermodynamic parameters were estimated using temperature-solubility product extrapolation and found to be logK S0 (Fe-Mc) = -34.59 ± 0.50, logK S0 (Fe-Hc) = -30.83 ± 0.50 and logK S0 (Al-Mc) = -31.32 ± 0.50.

  18. Dynamic behavior of RNA nanoparticles analyzed by AFM on a mica/air interface.

    Science.gov (United States)

    Sajja, Sameer; Chandler, Morgan; Federov, Dmitry; Kasprzak, Wojciech K; Lushnikov, Alexander Y; Viard, Mathias; Shah, Ankit; Dang, Dylan; Dahl, Jared; Worku, Beamlak; Dobrovolskaia, Marina A; Krasnoslobodtsev, Alexey; Shapiro, Bruce A; Afonin, Kirill A

    2018-04-18

    RNA is an attractive biopolymer for engineering self-assembling materials suitable for biomedical applications. Previously, programmable hexameric RNA rings were developed for the controlled delivery of up to six different functionalities. To increase the potential for functionalization with little impact on nanoparticle topology, we introduce gaps into the double-stranded regions of the RNA rings. Molecular dynamic simulations are used to assess the dynamic behavior and the changes in the flexibility of novel designs. The changes suggested by simulations, however, cannot be clearly confirmed by the conventional techniques such as non-denaturing polyacrylamide gel electrophoresis (native-PAGE) and dynamic light scattering (DLS). Also, an in vitro analysis in primary cultures of human peripheral blood mononuclear cells does not reveal any discrepancy in immunological recognition of new assemblies. To address these deficiencies, we introduce a computer-assisted quantification strategy. This strategy is based on an algorithmic atomic force microscopy (AFM)-resolved deformation analysis of the RNA nanoparticles studied on a mica/air interface. We validate this computational method by manual image analysis and fitting it to the simulation-predicted results. The presented nanoparticle modification strategy and subsequent AFM-based analysis are anticipated to provide a broad spectrum approach for the future development of nucleic acid-based nanotechnology.

  19. Magnetic domain structure investigation of Bi: YIG-thin films by combination of AFM and cantilever-based aperture SNOM

    International Nuclear Information System (INIS)

    Vysokikh, Yu E; Shevyakov, V I; Krasnoborodko, S Yu; Shelaev, A V; Prokopov, A R

    2016-01-01

    We present the results of magnetic domain structure investigation by combination of atomic force microscopy (AFM) and scanning near-field optical microscopy (SNOM). Special hollow-pyramid AFM cantilevers with aperture was used. This combination allows us use same probe for both topography and domain structure visualization of Bi -substituted ferrite garnet films of micro- and nano-meter thickness. Samples were excited through aperture by tightly focused linearly polarized laser beam. Magneto-optical effect rotates polarization of transmitted light depend on domain orientation. Visualization of magnetic domains was performed by detecting cross polarized component of transmitted light. SNOM allows to obtain high resolution magnetic domain image and prevent sample from any disturbance by magnetic probe. Same area SNOM and MFM images are presented. (paper)

  20. Structural characterization of Fe/Ag bilayers by RBS and AFM

    International Nuclear Information System (INIS)

    Tunyogi, A.; Tancziko, F.; Osvath, Z.; Paszti, F.

    2008-01-01

    Fe/Ag thin films are intensively investigated due to their special magnetic properties. Recently a deposition-order dependent asymmetric interface has been found. When iron is grown on silver, the interface is sharp, while the growth of Ag on Fe results in a long, low-energy tail of the Ag peak in the Rutherford backscattering spectrometry (RBS) spectra. The main purpose of this paper is to show that the low-energy Ag tail is caused by grain boundary diffusion, and that, when elevating the growing temperature of the Ag layer this effect becomes more significant. Two sets of polycrystalline and epitaxial Fe/Ag bilayers were prepared simultaneously onto Si(1 1 1) and MgO(1 0 0), respectively. The iron layers were grown at 250 deg. C and annealed at 450 deg. C in both sets, while the Ag layer was grown in the first set at room temperature (RT) and in the second set at 250 deg. C (HT). The sample composition, the interface sharpness and the quality of the epitaxy were studied by Rutherford backscattering spectrometry (RBS) combined with channeling effect. The surface morphology was determined by atomic force microscopy (AFM). RBS spectra show that in the case of RT samples the epitaxial MgO/Fe/Ag bilayer has sharp, well-defined interface, while for the polycrystalline Si/Fe/Ag sample the silver peak has a low-energy tail. Both the Fe and Ag peaks smeared out in the case of HT samples. AFM-images show that the RT samples have a continuous Ag layer, while the HT samples have fragmented surfaces. The RBS spectra taken on the HT samples were successfully simulated by the RBS-MAST code taking into account their fragmented structures.

  1. Methodological development of topographic correction in 2D/3D ToF-SIMS images using AFM images

    Science.gov (United States)

    Jung, Seokwon; Lee, Nodo; Choi, Myungshin; Lee, Jungmin; Cho, Eunkyunng; Joo, Minho

    2018-02-01

    Time-of-flight secondary-ion mass spectrometry (ToF-SIMS) is an emerging technique that provides chemical information directly from the surface of electronic materials, e.g. OLED and solar cell. It is very versatile and highly sensitive mass spectrometric technique that provides surface molecular information with their lateral distribution as a two-dimensional (2D) molecular image. Extending the usefulness of ToF-SIMS, a 3D molecular image can be generated by acquiring multiple 2D images in a stack. These imaging techniques by ToF-SIMS provide an insight into understanding the complex structures of unknown composition in electronic material. However, one drawback in ToF-SIMS is not able to represent topographical information in 2D and 3D mapping images. To overcome this technical limitation, topographic information by ex-situ technique such as atomic force microscopy (AFM) has been combined with chemical information from SIMS that provides both chemical and physical information in one image. The key to combine two different images obtained from ToF-SIMS and AFM techniques is to develop the image processing algorithm, which performs resize and alignment by comparing the specific pixel information of each image. In this work, we present methodological development of the semiautomatic alignment and the 3D structure interpolation system for the combination of 2D/3D images obtained by ToF-SIMS and AFM measurements, which allows providing useful analytical information in a single representation.

  2. Purification of nanoparticle PCR products and their topography observed with AFM

    International Nuclear Information System (INIS)

    Mi Lijuan; Wang Hubin; Chinese Academy of Sciences, Beijing; Li Bin; Zhou Hualan; Hu Jun

    2007-01-01

    Nanoparticle PCR (NP-PCR) is a new method to optimize PCR amplification. Suitable amount of Au nanoparticles can improve specificity, sensitivity and extension rate of PCR. In this paper, we compare efficiency of purifying NP-PCR products with different methods. In addition, topographies of DNA products in NP-PCR were observed with atomic force microscope (AFM). The results show that most of DNA products purified directly by routing method remain almost free due to less effect of nanoparticales. The yields decrease when the AuNPs were removed by high-speed centrifugation. A little amount of DNA subsided with AuNPs. (authors)

  3. A study on AFM manipulation of single-wall carbon nanotube

    Energy Technology Data Exchange (ETDEWEB)

    Tian Xiaojun; Dong Zaili; Yu Peng; Liu Zhu [State Key Lab. of Robotics, Shenyang Institute of Automation, Chinese Academy of Sciences, Shenyang 110016 (China)], E-mail: xjtian@sia.cn

    2009-09-01

    As single-wall carbon nanotube (SWCNT) has special electrical and physical property, it can be used as excellent material to construct various nano electronic device. However, in the fabrication process, the modification of size, shape and even the electronic property, especially to the metallic SWCNT, is a key problem to be overcome. Here a modified nanomanipulation technology based on atomic force microscope (AFM) is utilized to perform various kinds of SWCNT manipulation, such as SWCNT separation, catalyst remove, continual nano buckles fabrication and even stretch to break, thus to modify the size, shape and eventually the electrical property of the SWCNT. In addition, the manipulation results are analyzed based on the mechanical mechanism.

  4. An AFM and XPS study of corrosion caused by micro-liquid of dilute sulfuric acid on stainless steel

    International Nuclear Information System (INIS)

    Wang Rongguang

    2004-01-01

    Micro-liquid of dilute sulfuric acid deposited on SUS304 steel surface were observed with the ac non-contact mode of an atomic force microscopy (AFM), and the detail of the corrosion process caused by them was investigated with the contact mode of the AFM, X-ray photoelectron spectroscopy (XPS) and wavelength dispersive X-ray spectroscopy (WDXS). As a result, even not applying bias voltages between the tip of the cantilever and the specimen, micro-liquid of sulfuric acid can be successfully imaged using the ac non-contact mode of AFM. Two shapes of micro-acid, i.e., micro-droplets and micro-films, were found to co-exist on the specimen surface. On areas covered by micro-films of acid, only small corrosion product particles appeared and no corrosion pits were found. Beneath micro-droplets, corrosion reaction continue to produce pits until they were all consumed to form a corrosion product (mainly iron oxides) with almost the same shape with the droplet. The total corrosion reaction time was speculated to be between 690 and 1500 ks. The corrosion product formed from micro-droplets was believed to be a process of accumulating small corrosion product particles from the liquid/substrate interface to the surface of the formerly produced corrosion product. The XPS and WDXS analysis also supports the above results

  5. Coalescence and movement of nanobubbles studied with tapping mode AFM and tip-bubble interaction analysis

    International Nuclear Information System (INIS)

    Bhushan, Bharat; Wang Yuliang; Maali, Abdelhamid

    2008-01-01

    Imaging of a polystyrene (PS) coated silicon wafer immersed in deionized (DI) water was conducted using atomic force microscopy (AFM) in the tapping mode (TMAFM). As reported earlier, spherical cap-like domains, referred to as nanobubbles, were observed to be distributed on the PS surface. Experiments reveal that, in addition to the well-known parameter of scan load, scan speed is also an important parameter which affects nanobubble coalescence. The process of nanobubble coalescence was studied. It was found that during coalescence, small nanobubbles were easily moved and merged into bigger ones. Based on the interaction between the AFM cantilever tip and a bubble in the so-called force modulation mode of TMAFM, bubble height and adhesive force information for a given bubble was extracted. A viscoelastic model is used to obtain the interaction stiffness and damping coefficient, which provides a method to obtain the mechanical properties of nanobubbles. The model was further used to study the effect of surface tension force on attractive interaction force and contact angle hysteresis on the changes of the interaction damping coefficient during tip-bubble interaction.

  6. Large coercivity and unconventional exchange coupling in manganese-oxide-coated manganese—gallium nanoparticles

    International Nuclear Information System (INIS)

    Feng Jun-Ning; Liu Wei; Geng Dian-Yu; Ma Song; Yu Tao; Zhao Xiao-Tian; Dai Zhi-Ming; Zhao Xin-Guo; Zhang Zhi-Dong

    2014-01-01

    The microstructures and magnetic properties of nanoparticles, each composed of an antiferromagnetic (AFM) manganese-oxide shell and a ferromagnetic-like core of manganese—gallium (MnGa) compounds, are studied. The core-shell structure is confirmed by transmission electron microscope (TEM). The ferromagnetic-like core contains three kinds of MnGa binary compounds, i.e., ferrimagnetic (FI) D0 22 -type Mn 3 Ga, ferromagnetic (FM) Mn 8 Ga 5 , and AFM D0 19 -type Mn 3 Ga, of which the first two correspond respectively to a hard magnetic phase and to a soft one. Decoupling effect between these two phases is found at low temperature, which weakens gradually with increasing temperature and disappears above 200 K. The exchange bias (EB) effect is observed simultaneously, which is caused by the exchange coupling between the AFM shell and FM-like core. A large coercivity of 6.96 kOe (1 Oe = 79.5775 A·m −1 ) and a maximum EB value of 0.45 kOe are achieved at 300 K and 200 K respectively. (special topic — international conference on nanoscience and technology, china 2013)

  7. AFM and XRD characterization of silver nanoparticles films deposited on the surface of DGEBA epoxy resin by ion sputtering

    Energy Technology Data Exchange (ETDEWEB)

    Andrade, Jose Elisandro de; Machado, Rogerio; Macedo, Marcelo Andrade; Cunha, Frederico Guilherme Carvalho [Clinica de Medicina Nuclear e Radiologia de Maceio (MedRadiUS), Radiology and Imaging Diagnosis at Universidade Federal de Alagoas (UFAL), Maceio, AL (Brazil)

    2012-07-01

    In this work, silver atoms were deposited by ion sputtering on the surface of diglycidyl ether of bisphenol A (DGEBA) epoxy resin cured at 150 deg C for 6 hours in air. The films of DGEBA and its precursors were characterized by Raman spectroscopy to identify the main functional groups and their relationship with the deposited silver atoms. Silver thin films of 5, 10, 15 and 20 nm were deposited on the epoxy resin at room temperature. Both the initial film of DGEBA and the subsequent silver thin film were analyzed by Atomic Force Microscopy (AFM) in the non-contact mode. Silver thin films were also analyzed using X-ray diffraction (XRD) at room temperature. The AFM results showed the formation of silver crystallites on the surface of DGEBA at very low coverage whereas XRD indicated that most of them had their main axis aligned to the normal of the surface. An increase in the coverage led to an increase in the grain size as indicated by AFM. However, XRD results indicated that the crystallite size remained almost constant while the appearance of peaks corresponding to other crystalline orientations suggests the coalescence of the original crystallites and an increase in size of the more dense planes, namely [111]. (author)

  8. AFM and XRD characterization of silver nanoparticles films deposited on the surface of DGEBA epoxy resin by ion sputtering

    Energy Technology Data Exchange (ETDEWEB)

    Andrade, Jose Elisandro de; Machado, Rogerio; Macedo, Marcelo Andrade [Universidade Federal de Sergipe (UFSE), Aracaju, SE (Brazil). Programa de Pos-graduacao em Fisica; Cunha, Frederico Guilherme Carvalho [Universidade Federal de Sergipe (UFSE), Aracaju, SE (Brazil). Programa de Pos-graduacao em Ciencia e Engenharia de Materiais

    2012-07-01

    In this work, silver atoms were deposited by ion sputtering on the surface of diglycidyl ether of bisphenol A (DGEBA) epoxy resin cured at 150 Degree-Sign C for 6 hours in air. The films of DGEBA and its precursors were characterized by Raman spectroscopy to identify the main functional groups and their relationship with the deposited silver atoms. Silver thin films of 5, 10, 15 and 20 nm were deposited on the epoxy resin at room temperature. Both the initial film of DGEBA and the subsequent silver thin film were analyzed by Atomic Force Microscopy (AFM) in the non-contact mode. Silver thin films were also analyzed using X-ray diffraction (XRD) at room temperature. The AFM results showed the formation of silver crystallites on the surface of DGEBA at very low coverage whereas XRD indicated that most of them had their main axis aligned to the normal of the surface. An increase in the coverage led to an increase in the grain size as indicated by AFM. However, XRD results indicated that the crystallite size remained almost constant while the appearance of peaks corresponding to other crystalline orientations suggests the coalescence of the original crystallites and an increase in size of the more dense planes, namely [111]. (author)

  9. AFM and XRD characterization of silver nanoparticles films deposited on the surface of DGEBA epoxy resin by ion sputtering

    Directory of Open Access Journals (Sweden)

    José Elisandro de Andrade

    2013-01-01

    Full Text Available In this work, silver atoms were deposited by ion sputtering on the surface of diglycidyl ether of bisphenol A (DGEBA epoxy resin cured at 150 °C for 6 hours in air. The films of DGEBA and its precursors were characterized by Raman spectroscopy to identify the main functional groups and their relationship with the deposited silver atoms. Silver thin films of 5, 10, 15 and 20 nm were deposited on the epoxy resin at room temperature. Both the initial film of DGEBA and the subsequent silver thin film were analyzed by Atomic Force Microscopy (AFM in the non-contact mode. Silver thin films were also analyzed using X-ray diffraction (XRD at room temperature. The AFM results showed the formation of silver crystallites on the surface of DGEBA at very low coverage whereas XRD indicated that most of them had their main axis aligned to the normal of the surface. An increase in the coverage led to an increase in the grain size as indicated by AFM. However, XRD results indicated that the crystallite size remained almost constant while the appearance of peaks corresponding to other crystalline orientations suggests the coalescence of the original crystallites and an increase in size of the more dense planes, namely [111].

  10. Fabrication of novel AFM probe with high-aspect-ratio ultra-sharp three-face silicon nitride tips

    NARCIS (Netherlands)

    Vermeer, Rolf; Berenschot, Johan W.; Sarajlic, Edin; Tas, Niels Roelof; Jansen, Henricus V.

    In this paper we present the wafer-scale fabrication of molded AFM probes with high aspect ratio ultra-sharp three-plane silicon nitride tips. Using $\\langle$111$\\rangle$ silicon wafers a dedicated process is developed to fabricate molds in the silicon wafer that have a flat triangular bottom

  11. Deformation and wear of pyramidal, silicon-nitride AFM tips scanning micrometre-size features in contact mode

    NARCIS (Netherlands)

    Bloo, M.; Haitjema, H.; Pril, W.O.

    1999-01-01

    An experimental study was carried out, in order to investigate the deformation and wear taking place on pyramidal silicon-nitride AFM tips. The study focuses on the contact mode scanning of silicon features of micrometre-size. First the deformation and the mechanisms of wear of the tip during

  12. AFM studies of a new type of radiation defect on mica surfaces caused by highly charged ion impact

    International Nuclear Information System (INIS)

    Ruehlicke, C.; Briere, M.A.; Schneider, D.

    1994-01-01

    Radiation induced defects on mica caused by the impact of slow very highly charged ions (SVHCI) have been investigated with an atomic force microscope (AFM). Freshly cleaved surfaces of different types of muscovite were irradiated with SVHCI extracted from the LLNL electron beam ion trap (EBIT) at velocities of ca. 2 keV/amu. Atomic force microscopy of the surface reveals the formation of blisterlike defects associated with single ion impact. The determined defect volume which appears to increase linearly with the incident charge state and exhibits a threshold incident charge state has been determined using the AFM. These results indicate that target atoms are subjected to mutual electrostatic repulsion due to ionization through potential electron emission upon approach of the ion. If the repulsion leads to permanent atomic displacement, surface defects are formed

  13. Disordered strictly jammed binary sphere packings attain an anomalously large range of densities

    Science.gov (United States)

    Hopkins, Adam B.; Stillinger, Frank H.; Torquato, Salvatore

    2013-08-01

    Previous attempts to simulate disordered binary sphere packings have been limited in producing mechanically stable, isostatic packings across a broad spectrum of packing fractions. Here we report that disordered strictly jammed binary packings (packings that remain mechanically stable under general shear deformations and compressions) can be produced with an anomalously large range of average packing fractions 0.634≤ϕ≤0.829 for small to large sphere radius ratios α restricted to α≥0.100. Surprisingly, this range of average packing fractions is obtained for packings containing a subset of spheres (called the backbone) that are exactly strictly jammed, exactly isostatic, and also generated from random initial conditions. Additionally, the average packing fractions of these packings at certain α and small sphere relative number concentrations x approach those of the corresponding densest known ordered packings. These findings suggest for entropic reasons that these high-density disordered packings should be good glass formers and that they may be easy to prepare experimentally. We also identify an unusual feature of the packing fraction of jammed backbones (packings with rattlers excluded). The backbone packing fraction is about 0.624 over the majority of the α-x plane, even when large numbers of small spheres are present in the backbone. Over the (relatively small) area of the α-x plane where the backbone is not roughly constant, we find that backbone packing fractions range from about 0.606 to 0.829, with the volume of rattler spheres comprising between 1.6% and 26.9% of total sphere volume. To generate isostatic strictly jammed packings, we use an implementation of the Torquato-Jiao sequential linear programming algorithm [Phys. Rev. EPLEEE81539-375510.1103/PhysRevE.82.061302 82, 061302 (2010)], which is an efficient producer of inherent structures (mechanically stable configurations at the local maxima in the density landscape). The identification and

  14. Replication and dimensional quality control of industrial nanoscale surfaces using calibrated AFM measurements and SEM image processing

    DEFF Research Database (Denmark)

    Tosello, Guido; Hansen, Hans Nørgaard; Marinello, F.

    2010-01-01

    application of AFM to calibrate height, depth and pitch of sub-micrometer features and SEM image processing to detect replication accuracy in terms of number of replicated features. Surface replication is analyzed using a metrological approach: nano-features on nickel stampers and injection...

  15. A wideband large dynamic range and high linearity RF front-end for U-band mobile DTV

    International Nuclear Information System (INIS)

    Liu Rongjiang; Liu Shengyou; Guo Guiliang; Cheng Xu; Yan Yuepeng

    2013-01-01

    A wideband large dynamic range and high linearity U-band RF front-end for mobile DTV is introduced, and includes a noise-cancelling low-noise amplifier (LNA), an RF programmable gain amplifier (RFPGA) and a current communicating passive mixer. The noise/distortion cancelling structure and RC post-distortion compensation are employed to improve the linearity of the LNA. An RFPGA with five stages provides large dynamic range and fine gain resolution. A simple resistor voltage network in the passive mixer decreases the gate bias voltage of the mixing transistor, and optimum linearity and symmetrical mixing is obtained at the same time. The RF front-end is implemented in a 0.25 μm CMOS process. Tests show that it achieves an IIP3 (third-order intercept point) of −17 dBm, a conversion gain of 39 dB, and a noise figure of 5.8 dB. The RFPGA achieves a dynamic range of −36.2 to 23.5 dB with a resolution of 0.32 dB. (semiconductor integrated circuits)

  16. Real-Time Adaptive Control of a Magnetic Levitation System with a Large Range of Load Disturbance.

    Science.gov (United States)

    Zhang, Zhizhou; Li, Xiaolong

    2018-05-11

    In an idle light-load or a full-load condition, the change of the load mass of a suspension system is very significant. If the control parameters of conventional control methods remain unchanged, the suspension performance of the control system deteriorates rapidly or even loses stability when the load mass changes in a large range. In this paper, a real-time adaptive control method for a magnetic levitation system with large range of mass changes is proposed. First, the suspension control system model of the maglev train is built up, and the stability of the closed-loop system is analyzed. Then, a fast inner current-loop is used to simplify the design of the suspension control system, and an adaptive control method is put forward to ensure that the system is still in a stable state when the load mass varies in a wide range. Simulations and experiments show that when the load mass of the maglev system varies greatly, the adaptive control method is effective to suspend the system stably with a given displacement.

  17. AFM imaging reveals the tetrameric structure of the TRPC1 channel

    International Nuclear Information System (INIS)

    Barrera, Nelson P.; Shaifta, Yasin; McFadzean, Ian; Ward, Jeremy P.T.; Henderson, Robert M.; Edwardson, J. Michael

    2007-01-01

    We have determined the subunit stoichiometry of the transient receptor potential C1 (TRPC1) channel by imaging isolated channels using atomic force microscopy (AFM). A frequency distribution of the molecular volumes of individual channel particles had two peaks, at 170 and 720 nm 3 , corresponding with the expected sizes of TRPC1 monomers and tetramers, respectively. Complexes were formed between TRPC1 channels and antibodies against a V5 epitope tag present on each subunit. The frequency distribution of angles between pairs of bound antibodies had two peaks, at 88 o and 178 o . This result again indicates that the channel assembles as a tetramer

  18. Proceedings of the 2010 AFMS Medical Research Symposium. Volume 5. Nursing Track: Abstracts and Presentations

    Science.gov (United States)

    2011-03-15

    Management Education at a Military Hospital .................................................................. 29  Proceedings of the 2010 AFMS... education information is in Appendix C of this volume. Appendices D-L are copies of presentation slides from the plenary sessions.  Volume 2. This volume...women who have menstruation as compared to deployed women who do not have menstruation ? This study is a descriptive co relational research design. The

  19. Very large-scale structures in sintered silica aerogels as evidenced by atomic force microscopy and ultra-small angle X-ray scattering experiments

    CERN Document Server

    Marliere, C; Etienne, P; Woignier, T; Dieudonné, P; Phalippou, J

    2001-01-01

    During the last few years the bulk structure of silica aerogels has been extensively studied mainly by scattering techniques (neutrons, X-rays, light). It has been shown that small silica particles aggregate to constitute a fractal network. Its spatial extension and fractal dimension are strongly dependent on the synthesis conditions (e.g., pH of gelifying solutions). These typical lengths range from 1 to 10 nm. Ultra-small angle X-ray scattering (USAXS) and atomic force microscopy (AFM) experiments have been carried out on aerogels at different steps of densification. The results presented in this paper reveal the existence of a spatial arrangement of the solid part at a very large length scale. The evolution of this very large-scale structure during the densification process has been studied and reveals a contraction of this macro-structure made of aggregates of clusters. (16 refs).

  20. Atom probe, AFM and STM study on vacuum fired stainless steel

    International Nuclear Information System (INIS)

    Stupnik, A.; Frank, P.; Leisch, M.

    2008-01-01

    Full text: Stainless steel is one of the most commonly used structural materials for vacuum equipment. An efficient method to reduce the outgassing rate from stainless steel is a high temperature bakeout in vacuum (vacuum firing). This procedure reduces significantly the amount of dissolved hydrogen in the bulk. For the outgassing process the recombination rate of hydrogen atoms to the molecules plays the determining role and recombination is strongly related to the surface structure and composition. To get more detailed information about the surface morphology and composition AFM, STM and atom probe studies were carried out. Experiments on AISI 304L stainless steel samples show that the surface reconstructs completely during vacuum firing and large atomically flat terraces bounded by bunched steps and facets are formed. The large flat terraces can be assigned to (111) planes. The bunched steps and facets are corresponding in orientation almost to (110) planes and (100) planes. Surface inspection after vacuum firing by Auger electron spectroscopy (AES) gives reason for a composition change indicated by a reduction of the chromium signal in relation to the iron and nickel signal. Since the information depth of AES covers several atomic layers not only the top atomic layer of the sample surface is probed. For this reason 3D atom probe was used as well suited tool to investigate the segregation behavior of this alloy with the goal to examine the change in local chemical composition due to the high temperature treatment. As a result of vacuum firing the atom probe experiments show a significant enrichment of nickel at the top surface layer. In the second atomic layer chromium enrichment is detected. After vacuum firing the average composition below the second atomic layer shows certain chromium depletion up to 2 nm in depth. The observed changes in surface chemistry influence recombination and desorption probability from the surface and may contribute to the present

  1. Stiffness and evolution of interfacial micropancakes revealed by AFM quantitative nanomechanical imaging.

    Science.gov (United States)

    Zhao, Binyu; Wang, Xingya; Song, Yang; Hu, Jun; Lü, Junhong; Zhou, Xingfei; Tai, Renzhong; Zhang, Xuehua; Zhang, Lijuan

    2015-05-28

    Micropancakes are quasi-two-dimensional micron-sized domains on crystalline substrates (e.g. highly oriented pyrolytic graphite (HOPG)) immersed in water. They are only a few nanometers thick, and are suspected to come from the accumulation of dissolved air at the solid-water interface. However, the exact chemical nature and basic physical properties of micropancakes have been under debate ever since their first observation, primarily due to the lack of a suitable characterization technique. In this study, the stiffness of micropancakes at the interface between HOPG and ethanol-water solutions was investigated by using PeakForce Quantitative NanoMechanics (PF-QNM) mode Atomic Force Microscopy (AFM). Our measurements showed that micropancakes were stiffer than nanobubbles, and for bilayer micropancakes, the bottom layer in contact with the substrate was stiffer than the top one. Interestingly, the micropancakes became smaller and softer with an increase in the ethanol concentration in the solution, and were undetectable by AFM above a critical concentration of ethanol. But they re-appeared after the ethanol concentration in the solution was reduced. Clearly the evolution and stiffness of the micropancakes were dependent on the chemical composition in the solution, which could be attributed to the correlation of the mechanical properties of the micropancakes with the surface tension of the liquid phase. Based on the "go-and-come" behaviors of micropancakes with the ethanol concentration, we found that the micropancakes could actually tolerate the ethanol concentration much higher than 5%, a value reported in the literature. The results from this work may be helpful in alluding the chemical nature of micropancakes.

  2. Three-channel false colour AFM images for improved interpretation of complex surfaces: A study of filamentous cyanobacteria

    International Nuclear Information System (INIS)

    Kurk, Toby; Adams, David G.; Connell, Simon D.; Thomson, Neil H.

    2010-01-01

    Imaging signals derived from the atomic force microscope (AFM) are typically presented as separate adjacent images with greyscale or pseudo-colour palettes. We propose that information-rich false-colour composites are a useful means of presenting three-channel AFM image data. This method can aid the interpretation of complex surfaces and facilitate the perception of information that is convoluted across data channels. We illustrate this approach with images of filamentous cyanobacteria imaged in air and under aqueous buffer, using both deflection-modulation (contact) mode and amplitude-modulation (tapping) mode. Topography-dependent contrast in the error and tertiary signals aids the interpretation of the topography signal by contributing additional data, resulting in a more detailed image, and by showing variations in the probe-surface interaction. Moreover, topography-independent contrast and topography-dependent contrast in the tertiary data image (phase or friction) can be distinguished more easily as a consequence of the three dimensional colour-space.

  3. Three-channel false colour AFM images for improved interpretation of complex surfaces: A study of filamentous cyanobacteria

    Energy Technology Data Exchange (ETDEWEB)

    Kurk, Toby, E-mail: phytak@leeds.ac.uk [School of Physics and Astronomy, University of Leeds, Leeds, West Yorkshire LS2 9JT (United Kingdom); Adams, David G., E-mail: D.G.Adams@leeds.ac.uk [Faculty of Biological Sciences, University of Leeds, Leeds, West Yorkshire LS2 9JT (United Kingdom); Connell, Simon D., E-mail: S.D.A.Connell@leeds.ac.uk [School of Physics and Astronomy, University of Leeds, Leeds, West Yorkshire LS2 9JT (United Kingdom); Thomson, Neil H., E-mail: N.H.Thomson@leeds.ac.uk [School of Physics and Astronomy, University of Leeds, Leeds, West Yorkshire LS2 9JT (United Kingdom); Dental Institute, University of Leeds, Leeds, West Yorkshire LS2 9JT (United Kingdom)

    2010-05-15

    Imaging signals derived from the atomic force microscope (AFM) are typically presented as separate adjacent images with greyscale or pseudo-colour palettes. We propose that information-rich false-colour composites are a useful means of presenting three-channel AFM image data. This method can aid the interpretation of complex surfaces and facilitate the perception of information that is convoluted across data channels. We illustrate this approach with images of filamentous cyanobacteria imaged in air and under aqueous buffer, using both deflection-modulation (contact) mode and amplitude-modulation (tapping) mode. Topography-dependent contrast in the error and tertiary signals aids the interpretation of the topography signal by contributing additional data, resulting in a more detailed image, and by showing variations in the probe-surface interaction. Moreover, topography-independent contrast and topography-dependent contrast in the tertiary data image (phase or friction) can be distinguished more easily as a consequence of the three dimensional colour-space.

  4. Performance verification of Surface Mapping Instrument developed at CGM

    DEFF Research Database (Denmark)

    Bariani, Paolo

    The need of measuring narrow structures, in the micro and nano scale, over a broader range, can be satisfied by the use of highly resolving techniques, such as atomic force microscopy (AFM), in combination with probe relocation and data file stitching. At the Technical University of Denmark......, research has been carried out over the past years involving an AFM probe mounted on a coordinate measuring machine (CMM). Sensor repositioning by the CMM has made possible the inspection of relatively large samples, which are normally not investigable with AFMs. The latest step in the development...

  5. AFM-based nanolithography : manipulating poly(dimethylsiloxane) : loading force, scan speed and image resolution dependence on stick-slip outcomes in the slow and fast scan directions

    International Nuclear Information System (INIS)

    Watson, J.A.; Brown, C.L.; Myhra, S.; Watson, G.S.

    2005-01-01

    The various properties of a polymer will affect its functionality in a wide range of applications including biosensors, tissue engineering and biomaterials technology. Some of those require precise manipulation of laterally differentiated regions, currently taking place on the μm-scale. It is now apparent that this need must now be driven into the nm-regime. Using the AFM, the principal objective is to explore and investigate lithographic outcomes during tip-induced manipulation with the aid of work on poly(dimethylsiloxane), (PDMS). The frictional effects (including any in-plane relaxation), and their dependence on the loading force, scan speed and image resolution are examined. (author). 3 refs., 5 figs

  6. Optimization of functionalization conditions for protein analysis by AFM

    Energy Technology Data Exchange (ETDEWEB)

    Arroyo-Hernández, María, E-mail: maria.arroyo@ctb.upm.es [Centro de Tecnología Biomédica, Universidad Politécnica de Madrid, 28223 Pozuelo de Alarcón, Madrid (Spain); Departamento de Ciencia de Materiales, ETSI Caminos, Canales y Puertos, Universidad Politécnica de Madrid, 28040 Madrid (Spain); Daza, Rafael; Pérez-Rigueiro, Jose; Elices, Manuel; Nieto-Márquez, Jorge; Guinea, Gustavo V. [Centro de Tecnología Biomédica, Universidad Politécnica de Madrid, 28223 Pozuelo de Alarcón, Madrid (Spain); Departamento de Ciencia de Materiales, ETSI Caminos, Canales y Puertos, Universidad Politécnica de Madrid, 28040 Madrid (Spain)

    2014-10-30

    Highlights: • Highest fluorescence is obtained for central conditions. • Largest primary amine contribution is obtained for central conditions. • RMS roughness is smaller than 1 nm for all functional films. • Selected deposition conditions lead to proper RMS and functionality values. • LDH proteins adsorbed on AVS-films were observed by AFM. - Abstract: Activated vapor silanization (AVS) is used to functionalize silicon surfaces through deposition of amine-containing thin films. AVS combines vapor silanization and chemical vapor deposition techniques and allows the properties of the functionalized layers (thickness, amine concentration and topography) to be controlled by tuning the deposition conditions. An accurate characterization is performed to correlate the deposition conditions and functional-film properties. In particular, it is shown that smooth surfaces with a sufficient surface density of amine groups may be obtained with this technique. These surfaces are suitable for the study of proteins with atomic force microscopy.

  7. Effect of genomic long-range correlations on DNA persistence length: from theory to single molecule experiments.

    Science.gov (United States)

    Moukhtar, Julien; Faivre-Moskalenko, Cendrine; Milani, Pascale; Audit, Benjamin; Vaillant, Cedric; Fontaine, Emeline; Mongelard, Fabien; Lavorel, Guillaume; St-Jean, Philippe; Bouvet, Philippe; Argoul, Françoise; Arneodo, Alain

    2010-04-22

    Sequence dependency of DNA intrinsic bending properties has been emphasized as a possible key ingredient to in vivo chromatin organization. We use atomic force microscopy (AFM) in air and liquid to image intrinsically straight (synthetic), uncorrelated (hepatitis C RNA virus) and persistent long-range correlated (human) DNA fragments in various ionic conditions such that the molecules freely equilibrate on the mica surface before being captured in a particular conformation. 2D thermodynamic equilibrium is experimentally verified by a detailed statistical analysis of the Gaussian nature of the DNA bend angle fluctuations. We show that the worm-like chain (WLC) model, commonly used to describe the average conformation of long semiflexible polymers, reproduces remarkably well the persistence length estimates for the first two molecules as consistently obtained from (i) mean square end-to-end distance measurement and (ii) mean projection of the end-to-end vector on the initial orientation. Whatever the operating conditions (air or liquid, concentration of metal cations Mg(2+) and/or Ni(2+)), the persistence length found for the uncorrelated viral DNA underestimates the value obtained for the straight DNA. We show that this systematic difference is the signature of the presence of an uncorrelated structural intrinsic disorder in the hepatitis C virus (HCV) DNA fragment that superimposes on local curvatures induced by thermal fluctuations and that only the entropic disorder depends upon experimental conditions. In contrast, the WLC model fails to describe the human DNA conformations. We use a mean-field extension of the WLC model to account for the presence of long-range correlations (LRC) in the intrinsic curvature disorder of human genomic DNA: the stronger the LRC, the smaller the persistence length. The comparison of AFM imaging of human DNA with LRC DNA simulations confirms that the rather small mean square end-to-end distance observed, particularly for G

  8. BOREAS AFM-08 ECMWF Hourly Surface and Upper Air Data for the SSA and NSA

    Science.gov (United States)

    Viterbo, Pedro; Betts, Alan; Hall, Forrest G. (Editor); Newcomer, Jeffrey A.; Smith, David E. (Technical Monitor)

    2000-01-01

    The Boreal Ecosystem-Atmosphere Study (BOREAS) Airborne Fluxes and Meteorology (AFM)-8 team focused on modeling efforts to improve the understanding of the diurnal evolution of the convective boundary layer over the boreal forest. This data set contains hourly data from the European Center for for Medium-Range Weather Forecasts (ECMWF) operational model from below the surface to the top of the atmosphere, including the model fluxes at the surface. Spatially, the data cover a pair of the points that enclose the rawinsonde sites at Candle Lake, Saskatchewan, in the Southern Study Area (SSA) and Thompson, Manitoba, in the Northern Study Area (NSA). Temporally, the data include the two time periods of 13 May 1994 to 30 Sept 1994 and 01 Mar 1996 to 31 Mar 1997. The data are stored in tabular ASCII files. The number of records in the upper air data files may exceed 20,000, causing a problem for some software packages. The ECMWF hourly surface and upper air data are available from the Earth Observing System Data and Information System (EOSDIS) Oak Ridge National Laboratory (ORNL) Distributed Active Archive Center (DAAC). The data files are available on a CD-ROM (see document number 20010000884).

  9. Mechanical properties of in situ demineralised human enamel measured by AFM nanoindentation

    Science.gov (United States)

    Finke, Manuela; Hughes, Julie A.; Parker, David M.; Jandt, Klaus D.

    2001-10-01

    Diet-induced demineralisation is one of the key factors in surface changes of tooth enamel, with soft drinks being a significant etiological agent. The first step in this dissolution process is characterised by a change in the mechanical properties of the enamel and a roughening of the surface. The objective of this pilot study was to measure early stages of in situ induced hardness changes of polished human enamel surfaces with high accuracy using a nanoindenter attached to an atomic force microscope (AFM). Human unerupted third molars were cleaned, sterilised with sodium hypochlorite, sectioned and embedded in epoxy resin. The outer enamel surface was polished and the samples partly covered with a tape, allowing a 2-mm-wide zone to be exposed to the oral environment. Samples were fitted in an intra-oral appliance, which was worn from 9 a.m. to 5 p.m. for one day. During this time the volunteer sipped 250 ml of a drink over 10 min periods at 9.00, 11.00, 13.00 and 15.00 h. Three different drinks, mineral water, orange juice and the prototype of a blackcurrant drink with low demineralisation potential were used in this study. At the end of the experiment the samples were detached from the appliance, the tape removed and the surfaces chemically cleaned. The surface hardness and reduced Young's modulus of the exposed and unexposed areas of each sample were determined. In addition, high resolution topographical AFM images were obtained. This study shows that by determining the hardness and reduced Young's modulus, the difference in demineralisation caused by the drinks can be detected and quantified before statistically significant changes in surface topography could be observed with the AFM. The maximum decrease in surface hardness and Young's modulus occurred in the samples exposed to orange juice, followed by those exposed to the blackcurrant drink, while exposure to water led to the same values as unexposed areas. A one-way ANOVA showed a statistically significant

  10. Large scale metal-free synthesis of graphene on sapphire and transfer-free device fabrication.

    Science.gov (United States)

    Song, Hyun Jae; Son, Minhyeok; Park, Chibeom; Lim, Hyunseob; Levendorf, Mark P; Tsen, Adam W; Park, Jiwoong; Choi, Hee Cheul

    2012-05-21

    Metal catalyst-free growth of large scale single layer graphene film on a sapphire substrate by a chemical vapor deposition (CVD) process at 950 °C is demonstrated. A top-gated graphene field effect transistor (FET) device is successfully fabricated without any transfer process. The detailed growth process is investigated by the atomic force microscopy (AFM) studies.

  11. Boundary driven Kawasaki process with long-range interaction: dynamical large deviations and steady states

    International Nuclear Information System (INIS)

    Mourragui, Mustapha; Orlandi, Enza

    2013-01-01

    A particle system with a single locally-conserved field (density) in a bounded interval with different densities maintained at the two endpoints of the interval is under study here. The particles interact in the bulk through a long-range potential parametrized by β⩾0 and evolve according to an exclusion rule. It is shown that the empirical particle density under the diffusive scaling solves a quasilinear integro-differential evolution equation with Dirichlet boundary conditions. The associated dynamical large deviation principle is proved. Furthermore, when β is small enough, it is also demonstrated that the empirical particle density obeys a law of large numbers with respect to the stationary measures (hydrostatic). The macroscopic particle density solves a non-local, stationary, transport equation. (paper)

  12. In situ QCM and TM-AFM investigations of the early stages of degradation of silver and copper surfaces

    International Nuclear Information System (INIS)

    Kleber, Ch.; Hilfrich, U.; Schreiner, M.

    2007-01-01

    The early stages of atmospheric corrosion of pure copper and pure silver specimens were investigated performing in situ tapping mode atomic force microscopy (TM-AFM), in situ quartz crystal microbalance (QCM) and X-ray photoelectron spectroscopy (XPS). The information obtained by TM-AFM is the change of the topography of the sample surfaces with emphasis on the shape and lateral distribution of the corrosion products grown within the first hours of weathering. The simultaneously performed in situ QCM measurements are indicating the mass changes due to possibly occurring corrosive processes on the surface during weathering and are therefore a valuable tool for the determination of corrosion rates. Investigations were carried out in synthetic air at different levels of relative humidity (RH) with and without addition of 250 ppb SO 2 as acidifying agent. On a polished copper surface the growth of corrosion products could be observed by TM-AFM analysis at 60% RH without any addition of acidifying gases [M. Wadsak, M. Schreiner, T. Aastrup, C. Leygraf, Surf. Sci. 454-456 (2000) 246-250]. On a weathered copper surface the addition of SO 2 to the moist air stream leads to the formation of additional features as already described in the literature [M. Wadsak, M. Schreiner, T. Aastrup, C. Leygraf, Surf. Sci. 454-456 (2000) 246-250; Ch. Kleber, J. Weissenrieder, M. Schreiner, C. Leygraf, Appl. Surf. Sci. 193 (2002) 245-253]. Exposing a silver specimen to humidity leads to the degradation of the surface structure as well as to a formation of corrosion products, which could be detected by in situ QCM measurements. After addition of 250 ppb SO 2 to the moist gas stream an increase of the formed feature's volume on the silver surface could be observed by TM-AFM measurements. The results obtained additionally from the in situ QCM measurements confirm the influence of SO 2 due to a further increase of the mass of the formed corrosion layer (and therefore an increase of the

  13. Removal of industrial dyes and heavy metals by Beauveria bassiana: FTIR, SEM, TEM and AFM investigations with Pb(II).

    Science.gov (United States)

    Gola, Deepak; Malik, Anushree; Namburath, Maneesh; Ahammad, Shaikh Ziauddin

    2017-10-01

    Presence of industrial dyes and heavy metal as a contaminant in environment poses a great risk to human health. In order to develop a potential technology for remediation of dyes (Reactive remazol red, Yellow 3RS, Indanthrene blue and Vat novatic grey) and heavy metal [Cu(II), Ni(II), Cd(II), Zn(II), Cr(VI) and Pb(II)] contamination, present study was performed with entomopathogenic fungi, Beauveria bassiana (MTCC no. 4580). High dye removal (88-97%) was observed during the growth of B. bassiana while removal percentage for heavy metals ranged from 58 to 75%. Further, detailed investigations were performed with Pb(II) in terms of growth kinetics, effect of process parameters and mechanism of removal. Growth rate decreased from 0.118 h -1 (control) to 0.031 h -1 , showing 28% reduction in biomass at 30 mg L -1 Pb(II) with 58.4% metal removal. Maximum Pb(II) removal was observed at 30 °C, neutral pH and 30 mg L -1 initial metal concentration. FTIR analysis indicated the changes induced by Pb(II) in functional groups on biomass surface. Further, microscopic analysis (SEM and atomic force microscopy (AFM)) was performed to understand the changes in cell surface morphology of the fungal cell. SEM micrograph showed a clear deformation of fungal hyphae, whereas AFM studies proved the increase in surface roughness (RSM) in comparison to control cell. Homogenous bioaccumulation of Pb(II) inside the fungal cell was clearly depicted by TEM-high-angle annular dark field coupled with EDX. Present study provides an insight into the mechanism of Pb(II) bioremediation and strengthens the significance of using entomopathogenic fungus such as B. bassiana for metal and dye removal.

  14. Strategies for the AFM-based manipulation of silver nanowires on a flat surface

    Science.gov (United States)

    Liu, Hong-Zhi; Wu, Sen; Zhang, Jun-Ming; Bai, Hui-Tian; Jin, Fan; Pang, Hai; Hu, Xiao-Dong

    2017-09-01

    Silver nanowires (Ag NWs) are a promising material for building various sensors and devices at the nanoscale. However, the fast and precise placement of individual Ag NWs is still a challenge today. Atomic force microscopy (AFM) has been widely used to manipulate nanoparticles, yet this technology encounters many difficulties when being applied to the movement of Ag NWs as well as other soft one-dimensional (1D) materials, since the samples are easily distorted or even broken due to friction and adhesion on the substrate. In this paper, two novel manipulation strategies based on the parallel pushing method are presented. This method applies a group of short parallel pushing vectors (PPVs) to the Ag NW along its longitudinal direction. Identical and proportional vectors are respectively proposed to translate and rotate the Ag NWs with a straight-line configuration. The rotation strategy is also applied to straighten flexed Ag NWs. The finite element method simulation is introduced to analyse the behaviour of the Ag NWs as well as to optimize the parameter setting of the PPVs. Experiments are carried out to confirm the efficiency of the presented strategies. By comprehensive application of the new strategies, four Ag NWs are continuously assembled in a rectangular pattern. This study improves the controllability of the position and configuration of Ag NWs on a flat substrate. It also indicates the practicability of automatic nanofabrication using common AFMs.

  15. Influence of the tip mass on the tip-sample interactions in TM-AFM

    Energy Technology Data Exchange (ETDEWEB)

    Pishkenari, Hossein Nejat, E-mail: nejat@mech.sharif.edu [Nano-Robotics Laboratory, Center of Excellence in Design, Robotics and Automation, School of Mechanical Engineering, Sharif University of Technology, Tehran, P.O. Box 11365-9465 (Iran, Islamic Republic of); Meghdari, Ali [Nano-Robotics Laboratory, Center of Excellence in Design, Robotics and Automation, School of Mechanical Engineering, Sharif University of Technology, Tehran, P.O. Box 11365-9465 (Iran, Islamic Republic of)

    2011-07-15

    This paper focuses on the influences of the tip mass ratio (the ratio of the tip mass to the cantilever mass), on the excitation of higher oscillation eigenmodes and also on the tip-sample interaction forces in tapping mode atomic force microscopy (TM-AFM). A precise model for the cantilever dynamics capable of accurate simulations is essential for the investigation of the tip mass effects on the interaction forces. In the present work, the finite element method (FEM) is used for modeling the AFM cantilever to consider the oscillations of higher eigenmodes oscillations. In addition, molecular dynamics (MD) is used to calculate precise data for the tip-sample force as a function of tip vertical position with respect to the sample. The results demonstrate that in the presence of nonlinear tip-sample interaction forces, the tip mass ratio plays a significant role in the excitations of higher eigenmodes and also in the normal force applied on the surface. Furthermore, it has been shown that the difference between responses of the FEM and point-mass models in different system operational conditions is highly affected by the tip mass ratio. -- Highlights: {yields} A strong correlation exists between the tip mass ratio and the 18th harmonic amplitude. {yields} Near the critical tip mass ratio a small change in the tip mass may lead to a significant force change. {yields} Inaccuracy of the lumped model depends significantly on the tip mass ratio.

  16. Characterization of the structure of the coating of multilayers using AFM and Interferometric Microscopy

    International Nuclear Information System (INIS)

    Jerez A, Martha I; Lara O, Laura; Morantes M, Luz D; Plata G, Arturo; Torres, Yezid; Tsygankov, Petr

    2011-01-01

    Ti / TiN films were deposited on H13 steel and silicon substrates with different deposition voltage, by means of the cathodic arc evaporation (CAE) technique, this process was carried out by nanolayers deposition, requiring a detailed survey on growth films, for the properties characterization such as grain size, thickness and roughness of the film was used the atomic force microscopy (AFM) techniques and Interferometric Microscopy. Obtaining a the films growth when varying the deposition voltage.

  17. Improvement of the Correlative AFM and ToF-SIMS Approach Using an Empirical Sputter Model for 3D Chemical Characterization.

    Science.gov (United States)

    Terlier, T; Lee, J; Lee, K; Lee, Y

    2018-02-06

    Technological progress has spurred the development of increasingly sophisticated analytical devices. The full characterization of structures in terms of sample volume and composition is now highly complex. Here, a highly improved solution for 3D characterization of samples, based on an advanced method for 3D data correction, is proposed. Traditionally, secondary ion mass spectrometry (SIMS) provides the chemical distribution of sample surfaces. Combining successive sputtering with 2D surface projections enables a 3D volume rendering to be generated. However, surface topography can distort the volume rendering by necessitating the projection of a nonflat surface onto a planar image. Moreover, the sputtering is highly dependent on the probed material. Local variation of composition affects the sputter yield and the beam-induced roughness, which in turn alters the 3D render. To circumvent these drawbacks, the correlation of atomic force microscopy (AFM) with SIMS has been proposed in previous studies as a solution for the 3D chemical characterization. To extend the applicability of this approach, we have developed a methodology using AFM-time-of-flight (ToF)-SIMS combined with an empirical sputter model, "dynamic-model-based volume correction", to universally correct 3D structures. First, the simulation of 3D structures highlighted the great advantages of this new approach compared with classical methods. Then, we explored the applicability of this new correction to two types of samples, a patterned metallic multilayer and a diblock copolymer film presenting surface asperities. In both cases, the dynamic-model-based volume correction produced an accurate 3D reconstruction of the sample volume and composition. The combination of AFM-SIMS with the dynamic-model-based volume correction improves the understanding of the surface characteristics. Beyond the useful 3D chemical information provided by dynamic-model-based volume correction, the approach permits us to enhance

  18. Determination of Morphological Parameters of Supported Gold Nanoparticles: Comparison of AFM Combined with Optical Spectroscopy and Theoretical Modeling versus TEM

    Directory of Open Access Journals (Sweden)

    Frank Hubenthal

    2012-07-01

    Full Text Available The morphology of small gold particles prepared by Volmer–Weber growth on sapphire substrates have been investigated by two different characterization techniques. First, by non-extensive atomic force microscopy (AFM in combination with optical spectroscopy and modeling of the optical properties using a theoretical model, recently developed in our group. Second, by extensive transmission electron microscopy (TEM. Comparing the results obtained with both techniques demonstrate that for small gold nanoparticles within the quasistatic limit, the morphological properties can be precisely determined by an appropriate theoretical modeling of the optical properties in combination with simple AFM measurements. The apparent mean axial ratio of the nanoparticles, i.e., the axial ratio that corresponds to the center frequency of the ensemble plasmon resonance, is obtained easily from the extinction spectrum. The mean size is determined by the nanoparticle number density and the amount of deposited material, measured by AFM and a quartz micro balance, respectively. To extract the most probable axial ratio of the nanoparticle ensemble, i.e., the axial ratio that corresponds to the most probable nanoparticle size in the ensemble, we apply the new theoretical model, which allows to extract the functional dependence of the nanoparticle shape on its size. The morphological parameters obtained with this procedure will be afterwards compared to extensive TEM measurements. The results obtained with both techniques yield excellent agreement. For example, the lateral dimensions of the nanoparticles after deposition of 15.2 × 1015 atoms/cm2 of gold has been compared. While a mean lateral diameter of (13 ± 2 nm has been extracted from AFM, optical spectroscopy and modeling, a value of (12 ± 2 nm is derived from TEM. The consistency of the results demonstrate the precision of our new model. Moreover, since our theoretical model allows to extract the functional

  19. A large dynamic range radiation-tolerant analog memory in a quarter- micron CMOS technology

    CERN Document Server

    Anelli, G; Rivetti, A

    2001-01-01

    An analog memory prototype containing 8*128 cells has been designed in a commercial quarter-micron CMOS process. The aim of this work is to investigate the possibility of designing large dynamic range mixed-mode switched capacitor circuits for high-energy physics (HEP) applications in deep submicron CMOS technologies. Special layout techniques have been used to make the circuit radiation tolerant. The memory cells employ gate-oxide capacitors for storage, permitting a very high density. A voltage write-voltage read architecture has been chosen to minimize the sensitivity to absolute capacitor values. The measured input voltage range is 2.3 V (the power supply voltage V/sub DD/ is equal to 2.5 V), with a linearity of almost 8 bits over 2 V. The dynamic range is more than 11 bits. The pedestal variation is +or-0.5 mV peak-to-peak. The noise measured, which is dominated by the noise of the measurement setup, is around 0.8 mV rms. The characteristics of the memory have been measured before irradiation and after 1...

  20. A large dynamic range radiation tolerant analog memory in a quarter micron CMOS technology

    CERN Document Server

    Anelli, G; Rivetti, A

    2000-01-01

    A 8*128 cell analog memory prototype has been designed in a commercial 0.25 jam CMOS process. The aim of this work was to investigate the possibility of designing large dynamic range mixed- mode switched capacitor circuits for High-Energy Physics (HEP) applications in deep submicron CMOS technologies. Special layout techniques have been used to make the circuit radiation tolerant left bracket 1 right bracket . The memory cells employ gate-oxide capacitors for storage, allowing for a very high density. A voltage write - voltage read architecture has been chosen to minimize the sensitivity to absolute capacitor values. The measured input voltage range is 2.3 V (V//D//D = 2.5 V), with a linearity of at least 7.5 bits over 2 V. The dynamic range is more than 11 bits. The pedestal variation is plus or minus 0.5 mV peak-to-peak. The noise measured, which is dominated by the noise of the measurement setup, is around 0.8 mV rms. The characteristics of the memory have been measured before irradiation and after lOMrd (...

  1. Structure, cell wall elasticity and polysaccharide properties of living yeast cells, as probed by AFM

    International Nuclear Information System (INIS)

    Alsteens, David; Dupres, Vincent; Evoy, Kevin Mc; Dufrene, Yves F; Wildling, Linda; Gruber, Hermann J

    2008-01-01

    Although the chemical composition of yeast cell walls is known, the organization, assembly, and interactions of the various macromolecules remain poorly understood. Here, we used in situ atomic force microscopy (AFM) in three different modes to probe the ultrastructure, cell wall elasticity and polymer properties of two brewing yeast strains, i.e. Saccharomyces carlsbergensis and S. cerevisiae. Topographic images of the two strains revealed smooth and homogeneous cell surfaces, and the presence of circular bud scars on dividing cells. Nanomechanical measurements demonstrated that the cell wall elasticity of S. carlsbergensis is homogeneous. By contrast, the bud scar of S. cerevisiae was found to be stiffer than the cell wall, presumably due to the accumulation of chitin. Notably, single molecule force spectroscopy with lectin-modified tips revealed major differences in polysaccharide properties of the two strains. Polysaccharides were clearly more extended on S. cerevisiae, suggesting that not only oligosaccharides, but also polypeptide chains of the mannoproteins were stretched. Consistent with earlier cell surface analyses, these findings may explain the very different aggregation properties of the two organisms. This study demonstrates the power of using multiple complementary AFM modalities for probing the organization and interactions of the various macromolecules of microbial cell walls

  2. Quantitative measurements of intercellular adhesion between a macrophage and cancer cells using a cup-attached AFM chip.

    Science.gov (United States)

    Kim, Hyonchol; Yamagishi, Ayana; Imaizumi, Miku; Onomura, Yui; Nagasaki, Akira; Miyagi, Yohei; Okada, Tomoko; Nakamura, Chikashi

    2017-07-01

    Intercellular adhesion between a macrophage and cancer cells was quantitatively measured using atomic force microscopy (AFM). Cup-shaped metal hemispheres were fabricated using polystyrene particles as a template, and a cup was attached to the apex of the AFM cantilever. The cup-attached AFM chip (cup-chip) approached a murine macrophage cell (J774.2), the cell was captured on the inner concave of the cup, and picked up by withdrawing the cup-chip from the substrate. The cell-attached chip was advanced towards a murine breast cancer cell (FP10SC2), and intercellular adhesion between the two cells was quantitatively measured. To compare cell adhesion strength, the work required to separate two adhered cells (separation work) was used as a parameter. Separation work was almost 2-fold larger between a J774.2 cell and FP10SC2 cell than between J774.2 cell and three additional different cancer cells (4T1E, MAT-LyLu, and U-2OS), two FP10SC2 cells, or two J774.2 cells. FP10SC2 was established from 4T1E as a highly metastatic cell line, indicates separation work increased as the malignancy of cancer cells became higher. One possible explanation of the strong adhesion of macrophages to cancer cells observed in this study is that the measurement condition mimicked the microenvironment of tumor-associated macrophages (TAMs) in vivo, and J774.2 cells strongly expressed CD204, which is a marker of TAMs. The results of the present study, which were obtained by measuring cell adhesion strength quantitatively, indicate that the fabricated cup-chip is a useful tool for measuring intercellular adhesion easily and quantitatively. Copyright © 2017 Elsevier B.V. All rights reserved.

  3. Investigation of scale effects and directionality dependence on friction and adhesion of human hair using AFM and macroscale friction test apparatus

    International Nuclear Information System (INIS)

    LaTorre, Carmen; Bhushan, Bharat

    2006-01-01

    Macroscale testing of human hair tribological properties has been widely used to aid in the development of better shampoos and conditioners. Recently, literature has focused on using the atomic force microscope (AFM) to study surface roughness, coefficient of friction, adhesive force, and wear (tribological properties) on the nanoscale in order to increase understanding about how shampoos and conditioners interact with the hair cuticle. Since there are both similarities and differences when comparing the tribological trends at both scales, it is thus recognized that scale effects are an important aspect of studying the tribology of hair. However, no microscale tribological data for hair exists in literature. This is unfortunate because many interactions between hair-skin, hair-comb, and hair-hair contact takes place at microasperities ranging from a few μm to hundreds of μm. Thus, to bridge the gap between the macro- and nanoscale data, as well as to gain a full understanding of the mechanisms behind the trends, it is now worthwhile to look at hair tribology on the microscale. Presented in this paper are coefficient of friction and adhesive force data on various scales for virgin and chemically damaged hair, both with and without conditioner treatment. Macroscale coefficient of friction was determined using a traditional friction test apparatus. Microscale and nanoscale tribological characterization was performed with AFM tips of various radii. The nano-, micro-, and macroscale trends are compared and the mechanisms behind the scale effects are discussed. Since the coefficient of friction changes drastically (on any scale) depending on whether the direction of motion is along or against the cuticle scales, the directionality dependence and responsible mechanisms are discussed

  4. AFM Morphology Study of Si1-Y GeY:H Films Deposited by LF PE CVD from Silane-Germane with Different

    International Nuclear Information System (INIS)

    Sanchez, L; Kosarev, A

    2005-01-01

    The morphology of Si 1-Y Ge Y :H films in the range of Y=0.23 to 0.9 has been studied by AFM. The films were deposited by Low Frequency (LF) PE CVD at substrate temperature T s =300 C and discharge frequency f=110 kHz from silane+germane mixture with and without, Ar and H 2 dilution. The films were deposited on silicon and glass substrates. AFM images were taken and analyzed for 2 x 2 mm 2 area. All the images demonstrated ''grain'' like structure, which was characterized by the height distribution function F(H) average roughness , standard height deviation Rq, lateral correlation length L c area distribution function F(s), mean grain area , diameter distribution function F(d), and mean grain diameter . The roughness of the films monotonically increases with Y for all dilutions, but more significantly in the films deposited without dilution. L c continuously grows with Y in the films deposited without dilution, while more complex behavior L c (Y) is observed in the films deposited with H- or Ar dilution. The sharpness of F(H) characterized by curtosis γ depends on dilution and the sharpest F(H) are for the films deposited with Ar (γ=5.30,Y=0.23) and without dilution (γ=4.3, Y=0.45). Isothermal annealing caused increase of , L c in the films deposited with H- and Ar dilutions, while in the films prepared without dilution the behavior was more complex, depending on the substrates. Significant narrowing of the height distribution was observed in the films deposited with H dilution or without dilution

  5. Single cell adhesion force measurement for cell viability identification using an AFM cantilever-based micro putter

    Science.gov (United States)

    Shen, Yajing; Nakajima, Masahiro; Kojima, Seiji; Homma, Michio; Kojima, Masaru; Fukuda, Toshio

    2011-11-01

    Fast and sensitive cell viability identification is a key point for single cell analysis. To address this issue, this paper reports a novel single cell viability identification method based on the measurement of single cell shear adhesion force using an atomic force microscopy (AFM) cantilever-based micro putter. Viable and nonviable yeast cells are prepared and put onto three kinds of substrate surfaces, i.e. tungsten probe, gold and ITO substrate surfaces. A micro putter is fabricated from the AFM cantilever by focused ion beam etching technique. The spring constant of the micro putter is calibrated using the nanomanipulation approach. The shear adhesion force between the single viable or nonviable cell and each substrate is measured using the micro putter based on the nanorobotic manipulation system inside an environmental scanning electron microscope. The adhesion force is calculated based on the deflection of the micro putter beam. The results show that the adhesion force of the viable cell to the substrate is much larger than that of the nonviable cell. This identification method is label free, fast, sensitive and can give quantitative results at the single cell level.

  6. Using AFM to probe the complexation of DNA with anionic lipids mediated by Ca(2+): the role of surface pressure.

    Science.gov (United States)

    Luque-Caballero, Germán; Martín-Molina, Alberto; Sánchez-Treviño, Alda Yadira; Rodríguez-Valverde, Miguel A; Cabrerizo-Vílchez, Miguel A; Maldonado-Valderrama, Julia

    2014-04-28

    Complexation of DNA with lipids is currently being developed as an alternative to classical vectors based on viruses. Most of the research to date focuses on cationic lipids owing to their spontaneous complexation with DNA. Nonetheless, recent investigations have revealed that cationic lipids induce a large number of adverse effects on DNA delivery. Precisely, the lower cytotoxicity of anionic lipids accounts for their use as a promising alternative. However, the complexation of DNA with anionic lipids (mediated by cations) is still in early stages and is not yet well understood. In order to explore the molecular mechanisms underlying the complexation of anionic lipids and DNA we proposed a combined methodology based on the surface pressure-area isotherms, Gibbs elasticity and Atomic Force Microscopy (AFM). These techniques allow elucidation of the role of the surface pressure in the complexation and visualization of the interfacial aggregates for the first time. We demonstrate that the DNA complexes with negatively charged model monolayers (DPPC/DPPS 4 : 1) only in the presence of Ca(2+), but is expelled at very high surface pressures. Also, according to the Gibbs elasticity plot, the complexation of lipids and DNA implies a whole fluidisation of the monolayer and a completely different phase transition map in the presence of DNA and Ca(2+). AFM imaging allows identification for the first time of specific morphologies associated with different packing densities. At low surface coverage, a branched net like structure is observed whereas at high surface pressure fibers formed of interfacial aggregates appear. In summary, Ca(2+) mediates the interaction between DNA and negatively charged lipids and also the conformation of the ternary system depends on the surface pressure. Such observations are important new generic features of the interaction between DNA and anionic lipids.

  7. Nano-dot and nano-pit fabrication on a GaAs substrate by a pulse applied AFM

    International Nuclear Information System (INIS)

    Kim, H C; Yu, J S; Ryu, S H

    2012-01-01

    The nano-patterning characteristics of GaAs is investigated using a pulse applied atomic force microscope (AFM). Very short range voltage pulses of micro to nano-seconds’ duration are applied to a conductive diamond-coated silicon (Si) tip in contact mode, to regulate the created feature size. The effects of pulse conditions such as pulse voltage, duration, frequency, offset voltage, anodization time, and applied tip pressure on nano-dot generation are characterized, based on the experiments. An interesting phenomenon, nano-pit creation instead of nano-dot creation, is observed when the applied pulse duration is less than 100 μs. Pulse frequency and offset voltage are also involved in nano-pit generation. The electrical spark discharge between the tip and the GaAs's surface is the most probable cause of the nano-pit creation and its generation mechanism is explained by considering the relevant pulse parameters. Nano-pits over 15 nm in depth are acquired on the GaAs substrate by adjusting the pulse conditions. This research facilitates the fabrication of more complex nano-structures on semiconductor materials since nano-dots and nano-pits could be easily made without any additional post-processes. (paper)

  8. Corroboration of Raman and AFM mapping to study Si nanocrystals embedded in SiO{sub 2}

    Energy Technology Data Exchange (ETDEWEB)

    Rani, Ekta, E-mail: ades.ekta@gmail.com [Laser Physics Applications Section, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Homi Bhabha National Institute, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Ingale, Alka A. [Laser Physics Applications Section, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Homi Bhabha National Institute, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Chaturvedi, A. [Laser Material Processing Division, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Joshi, M.P.; Kukreja, L.M. [Homi Bhabha National Institute, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India); Laser Material Processing Division, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India)

    2016-07-05

    Raman and atomic force microscopy (AFM) mapping on the same selected area are used to get unique information about the morphology of Si nanocrystals (NCs) embedded in SiO{sub 2}, which is difficult to obtain by any other conventional technique. The sensitivity of Raman spectroscopy to surface/interface and confinement effects in NCs is effectively used to correlate the Raman intensity profile in Raman mapping with the topography obtained from AFM to understand that Si NCs are clustered in i) smaller clusters (∼100 nm) organized closely in two dimensions (2D) and ii) big (∼2 μm) three dimensional (3D) isolated clusters, although the growth is carried out to be multilayer (Si/SiO{sub 2}). Raman mapping performed by varying the focal spot along the depth shows stacking of larger (>∼60 Å) to smaller sizes (<∼40 Å) Si NCs from bottom to top for some clusters. To understand the observed morphologies, further study of specially grown Si–SiO{sub 2} nanocomposites is performed, which suggest formation of smaller Si NCs at the top due to annealing at 800 °C in Si rich SiO{sub 2} and possible existence of thermal gradient in an insulating matrix of SiO{sub 2.} Larger Si NCs are formed in the laser induced plume (plasma) itself. - Graphical abstract: a) Schematic showing the expected stacking of Si NCs obtained from Raman mapping, performed by changing focal spot along the depth, b) top, c) middle and d) bottom region of the cluster. - Highlights: • Methodology is developed to obtain Raman and AFM mapping at same selected area. • To get unique information, difficult to obtain using other conventional techniques. • Clusters (∼100 nm–2 μm) of Si nanocrystals embedded in SiO{sub 2} matrix are formed. • Stacking of Si nanocrystals from bottom to top (10–1 nm) is observed in some clusters. • Stacking of Si nanocrystals is understood as due to annealing and thermal gradient.

  9. In situ AFM investigation of electrochemically induced surface-initiated atom-transfer radical polymerization.

    Science.gov (United States)

    Li, Bin; Yu, Bo; Zhou, Feng

    2013-02-12

    Electrochemically induced surface-initiated atom-transfer radical polymerization is traced by in situ AFM technology for the first time, which allows visualization of the polymer growth process. It affords a fundamental insight into the surface morphology and growth mechanism simultaneously. Using this technique, the polymerization kinetics of two model monomers were studied, namely the anionic 3-sulfopropyl methacrylate potassium salt (SPMA) and the cationic 2-(metharyloyloxy)ethyltrimethylammonium chloride (METAC). The growth of METAC is significantly improved by screening the ammonium cations by the addition of ionic liquid electrolyte in aqueous solution. Copyright © 2013 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  10. Changes in surface characteristics of two different resin composites after 1 year water storage: An SEM and AFM study.

    Science.gov (United States)

    Tekçe, Neslihan; Pala, Kansad; Demirci, Mustafa; Tuncer, Safa

    2016-11-01

    To evaluate changes in surface characteristics of two different resin composites after 1 year of water storage using a profilometer, Vickers hardness, scanning electron microscopy (SEM), and atomic force microscopy (AFM). A total of 46 composite disk specimens (10 mm in diameter and 2 mm thick) were fabricated using Clearfil Majesty Esthetic and Clearfil Majesty Posterior (Kuraray Medical Co, Tokyo, Japan). Ten specimens from each composite were used for surface roughness and microhardness tests (n = 10). For each composite, scanning electron microscope (SEM, n = 2) and atomic force microscope (AFM, n = 1) images were obtained after 24 h and 1 year of water storage. The data were analyzed using two-way analysis of variance and a post-hoc Bonferroni test. Microhardness values of Clearfil Majesty Esthetic decreased significantly (78.15-63.74, p = 0.015) and surface roughness values did not change after 1 year of water storage (0.36-0.39, p = 0.464). Clearfil Majesty Posterior microhardness values were quite stable (138.74-137.25, p = 0.784), and surface roughness values increased significantly (0.39-0.48, p = 0.028) over 1 year. One year of water storage caused microhardness values for Clearfil Majesty Esthetic to decrease and the surface roughness of Clearfil Majesty Posterior increased. AFM and SEM images demonstrated surface detoration of the materials after 1 year and ensured similar results with the quantitative test methods. SCANNING 38:694-700, 2016. © 2016 Wiley Periodicals, Inc. © Wiley Periodicals, Inc.

  11. Optical imaging beyond the diffraction limit by SNEM: effects of AFM tip modifications with thiol monolayers on imaging quality.

    Science.gov (United States)

    Cumurcu, Aysegul; Diaz, Jordi; Lindsay, Ian D; de Beer, Sissi; Duvigneau, Joost; Schön, Peter; Julius Vancso, G

    2015-03-01

    Tip-enhanced nanoscale optical imaging techniques such as apertureless scanning near-field optical microscopy (a-SNOM) and scanning near-field ellipsometric microscopy (SNEM) applications can suffer from a steady degradation in performance due to adhesion of atmospheric contaminants to the metal coated tip. Here, we demonstrate that a self-assembled monolayer (SAM) of ethanethiol (EtSH) is an effective means of protecting gold-coated atomic force microscopy (AFM) probe tips from accumulation of surface contaminants during prolonged exposure to ambient air. The period over which they yield consistent and reproducible results for scanning near-field ellipsometric microscopy (SNEM) imaging is thus extended. SNEM optical images of a microphase separated polystyrene-block-poly (methylmethacrylate) (PS-b-PMMA) diblock copolymer film, which were captured with bare and SAM-protected gold-coated AFM probes, both immediately after coating and following five days of storage in ambient air, were compared. During this period the intensity of the optical signals from the untreated gold tip fell by 66%, while those from the SAM protected tip fell by 14%. Additionally, gold coated AFM probe tips were modified with various lengths of alkanethiols to measure the change in intensity variation in the optical images with SAM layer thickness. The experimental results were compared to point dipole model calculations. While a SAM of 1-dodecanethiol (DoSH) was found to strongly suppress field enhancement we find that it can be locally removed from the tip apex by deforming the molecules under load, restoring SNEM image contrast. Copyright © 2014 Elsevier B.V. All rights reserved.

  12. Radiolysis of aqueous solutions of ammonium bicarbonate over a large dose range

    International Nuclear Information System (INIS)

    Draganic, Z.D.; Draganic, I.G.; Negron-Mendoza, A.; Navarro-Gonzales, R.; Albarran-Sanchez, M.G.; Sehested, K.

    1986-12-01

    0 2 -free aqueous solutions of 0.05 mol dm -3 ammonium bicarbonate were studied after receiving various doses of 60 Co gammas (0.001-170 Mrd) or krd pulses of 10 MeV electrons. Formate, oxalate, formaldehyde and an unidentified polymer (M w 14000-16000 daltons) were found to be the main radiolytic products. A large initial yield of formate in the γ-radiolysis, G(HCOO - ) = 2.2, is due to the reaction CO 2 - + HCO 3 - ↔ HC00 - + CO 3 - . The efficiency of organic synthesis within the large dose range studied is low and is explained by efficient pathways to the reformation of bicarbonate, where the reaction CO 2 - + CO 3 - is particurlarly significant. Computer fitting of the data obtained in the γ-radiolysis and by pulsed electron beam experiments gives k(CO 2 - + HCO 3 - ) = (2 ± 0.4)x10 3 dm 3 mol -1 s -1 , k(CO 2 - + CO 3 - ) = (5 ± 1)x10 7 dm 3 mol -1 s -1 , k(NH 2 + = HCO 3 - ) 4 dm 3 mol -1 s -1 and k(NH 2 + CO 3 - ) = (1.5 ± 0.5)x10 9 dm 3 mol -1 s -1 . (author)

  13. Range-Free Localization Schemes for Large Scale Sensor Networks

    National Research Council Canada - National Science Library

    He, Tian; Huang, Chengdu; Blum, Brain M; Stankovic, John A; Abdelzaher, Tarek

    2003-01-01

    .... Because coarse accuracy is sufficient for most sensor network applications, solutions in range-free localization are being pursued as a cost-effective alternative to more expensive range-based approaches...

  14. Fabrication of micro-dot arrays and micro-walls of acrylic acid/melamine resin on aluminum by AFM probe processing and electrophoretic coating

    International Nuclear Information System (INIS)

    Kurokawa, S.; Kikuchi, T.; Sakairi, M.; Takahashi, H.

    2008-01-01

    Micro-dot arrays and micro-walls of acrylic acid/melamine resin were fabricated on aluminum by anodizing, atomic force microscope (AFM) probe processing, and electrophoretic deposition. Barrier type anodic oxide films of 15 nm thickness were formed on aluminum and then the specimen was scratched with an AFM probe in a solution containing acrylic acid/melamine resin nano-particles to remove the anodic oxide film locally. After scratching, the specimen was anodically polarized to deposit acrylic acid/melamine resin electrophoretically at the film-removed area. The resin deposited on the specimen was finally cured by heating. It was found that scratching with the AFM probe on open circuit leads to the contamination of the probe with resin, due to positive shifts in the potential during scratching. Scratching of the specimen under potentiostatic conditions at -1.0 V, however, resulted in successful resin deposition at the film-removed area without probe contamination. The rate of resin deposition increased as the specimen potential becomes more positive during electrophoretic deposition. Arrays of resin dots with a few to several tens μm diameter and 100-1000 nm height, and resin walls with 100-1000 nm height and 1 μm width were obtained on specimens by successive anodizing, probe processing, and electrophoretic deposition

  15. Fabrication of micro-dot arrays and micro-walls of acrylic acid/melamine resin on aluminum by AFM probe processing and electrophoretic coating

    Energy Technology Data Exchange (ETDEWEB)

    Kurokawa, S.; Kikuchi, T.; Sakairi, M. [Graduate School of Engineering, Hokkaido University, N-13, W-8, Kita-Ku, Sapporo 060-8628 (Japan); Takahashi, H. [Graduate School of Engineering, Hokkaido University, N-13, W-8, Kita-Ku, Sapporo 060-8628 (Japan)], E-mail: takahasi@elechem1-mc.eng.hokudai.ac.jp

    2008-11-30

    Micro-dot arrays and micro-walls of acrylic acid/melamine resin were fabricated on aluminum by anodizing, atomic force microscope (AFM) probe processing, and electrophoretic deposition. Barrier type anodic oxide films of 15 nm thickness were formed on aluminum and then the specimen was scratched with an AFM probe in a solution containing acrylic acid/melamine resin nano-particles to remove the anodic oxide film locally. After scratching, the specimen was anodically polarized to deposit acrylic acid/melamine resin electrophoretically at the film-removed area. The resin deposited on the specimen was finally cured by heating. It was found that scratching with the AFM probe on open circuit leads to the contamination of the probe with resin, due to positive shifts in the potential during scratching. Scratching of the specimen under potentiostatic conditions at -1.0 V, however, resulted in successful resin deposition at the film-removed area without probe contamination. The rate of resin deposition increased as the specimen potential becomes more positive during electrophoretic deposition. Arrays of resin dots with a few to several tens {mu}m diameter and 100-1000 nm height, and resin walls with 100-1000 nm height and 1 {mu}m width were obtained on specimens by successive anodizing, probe processing, and electrophoretic deposition.

  16. Adsorption mechanisms for fatty acids on DLC and steel studied by AFM and tribological experiments

    International Nuclear Information System (INIS)

    Simič, R.; Kalin, M.

    2013-01-01

    Fatty acids are known to affect the friction and wear of steel contacts via adsorption onto the surface, which is one of the fundamental boundary-lubrication mechanisms. The understanding of the lubrication mechanisms of polar molecules on diamond-like carbon (DLC) is, however, still insufficient. In this work we aimed to find out whether such molecules have a similar effect on DLC coatings as they do on steel. The adsorption of hexadecanoic acid in various concentrations (2–20 mmol/l) on DLC was studied under static conditions using an atomic force microscope (AFM). The amount of surface coverage of the adsorbed fatty-acid molecules was analysed. In addition, tribological tests were performed to correlate the wear and friction behaviours in tribological contacts with the adsorption of molecules on the surface under static conditions. A good correlation between the AFM results and the tribological behaviour was observed. We confirmed that fatty acids can adsorb onto the DLC surfaces and are, therefore, potential boundary-lubrication agents for DLC coatings. The adsorption of the fatty acid onto the DLC surfaces reduces the wear of the coatings, but it is less effective in reducing the friction. Tentative adsorption mechanisms that include an environmental species effect, a temperature effect and a tribochemical effect are proposed for DLC and steel surfaces based on our results and few potential mechanisms found in literature.

  17. Adsorption mechanisms for fatty acids on DLC and steel studied by AFM and tribological experiments

    Energy Technology Data Exchange (ETDEWEB)

    Simič, R.; Kalin, M., E-mail: mitjan.kalin@tint.fs.uni-lj.si

    2013-10-15

    Fatty acids are known to affect the friction and wear of steel contacts via adsorption onto the surface, which is one of the fundamental boundary-lubrication mechanisms. The understanding of the lubrication mechanisms of polar molecules on diamond-like carbon (DLC) is, however, still insufficient. In this work we aimed to find out whether such molecules have a similar effect on DLC coatings as they do on steel. The adsorption of hexadecanoic acid in various concentrations (2–20 mmol/l) on DLC was studied under static conditions using an atomic force microscope (AFM). The amount of surface coverage of the adsorbed fatty-acid molecules was analysed. In addition, tribological tests were performed to correlate the wear and friction behaviours in tribological contacts with the adsorption of molecules on the surface under static conditions. A good correlation between the AFM results and the tribological behaviour was observed. We confirmed that fatty acids can adsorb onto the DLC surfaces and are, therefore, potential boundary-lubrication agents for DLC coatings. The adsorption of the fatty acid onto the DLC surfaces reduces the wear of the coatings, but it is less effective in reducing the friction. Tentative adsorption mechanisms that include an environmental species effect, a temperature effect and a tribochemical effect are proposed for DLC and steel surfaces based on our results and few potential mechanisms found in literature.

  18. Tip-surface interactions at redox responsive poly(ferrocenylsilane) (PFS) interface by AFM-based force spectroscopy

    International Nuclear Information System (INIS)

    Chung Hongjing; Song Jing; Vancso, G. Julius

    2009-01-01

    Poly(ferrocenylsilanes) (PFS) belong to the class of redox responsive organometallic polymers. Atomic force microscopy (AFM)-based single molecule force spectroscopy (SMFS) was used earlier to study single chain PFS response and redox energy driven single chain PFS molecular motors. Here we present further AFM investigations of force interactions between tip and a grafted PFS surface under potential control in electrochemical redox cycles. Typical tip-Au interaction is considered as reference in the force measurements. First the electrostatic component in the diffused double layer (DL) in NaClO 4 electrolyte environment was considered for a 'grafted to' PFS, which dominated the interplay between the tip and sample surface. The DL forces can also hinder the physisorption of PFS chain onto the tip when the voltage was applied at -0.1 V. On the other hand, if the tip contacted the PFS surface prior to the electrochemical process, physisorption of PFS chains governed the overall interaction regardless of subsequently applied surface potential. In addition, prolonged contact time, t c , may also contribute to the stability of tip-PFS bridging and detection of electrostatic forces between the tip-PFS interface. The results showed that tip-substrate interaction forces without PFS grafts have negligibly small force contributions under similar, electrochemically controlled, conditions used in single PFS chain based molecular motors.

  19. Surface-roughness-assisted formation of large-scale vertically aligned CdS nanorod arrays via solvothermal method

    Energy Technology Data Exchange (ETDEWEB)

    Zhou, Minmin [National Laboratory of Solid State Microstructures, School of Electronic Science and Engineering, Nanjing University, Nanjing 210093 (China); Yan, Shancheng, E-mail: yansc@njupt.edu.cn [National Laboratory of Solid State Microstructures, School of Electronic Science and Engineering, Nanjing University, Nanjing 210093 (China); School of Geography and Biological Information, Nanjing University of Posts and Telecommunications, Nanjing 210046 (China); Shi, Yi, E-mail: yshi@nju.edu.cn [National Laboratory of Solid State Microstructures, School of Electronic Science and Engineering, Nanjing University, Nanjing 210093 (China); Yang, Meng; Sun, Huabin; Wang, Jianyu; Yin, Yao; Gao, Fan [National Laboratory of Solid State Microstructures, School of Electronic Science and Engineering, Nanjing University, Nanjing 210093 (China)

    2013-05-15

    Large-scale cadmium sulfide (CdS) nanorod arrays were successfully synthesized on several different substrates through solvothermal reaction. During the growth experiments, we observed that the adhesion strength of the CdS nanorod arrays to different substrates differed dramatically, causing some of the CdS coating being easily flushed away by deionized water (DI water). With doubts and suspicions, we seriously investigate the original morphology of all the substrates by using atomic force microscopy (AFM). The phase, morphology, crystal structure and photoelectric property of all the products were characterized by X-ray diffractometer (XRD), scanning electron microscopy (SEM), transmission electron microscopy (TEM), Raman spectroscopy and current–voltage (I–V) probe station. The growth mechanism of solvothermal reaction was proposed on the basis of all the characterizations. Our approach presents a universal method of liquid phase epitaxy of 1D material on a wide range of substrates of any shape.

  20. AFM and SFG studies of pHEMA-based hydrogel contact lens surfaces in saline solution: adhesion, friction, and the presence of non-crosslinked polymer chains at the surface.

    Science.gov (United States)

    Kim, Seong Han; Opdahl, Aric; Marmo, Chris; Somorjai, Gabor A

    2002-04-01

    The surfaces of two types of soft contact lenses neutral and ionic hydrogels--were characterized by atomic force microscopy (AFM) and sum-frequency-generation (SFG) vibrational spectroscopy. AFM measurements in saline solution showed that the presence of ionic functional groups at the surface lowered the friction and adhesion to a hydrophobic polystyrene tip. This was attributed to the specific interactions of water and the molecular orientation of hydrogel chains at the surface. Friction and adhesion behavior also revealed the presence of domains of non-crosslinked polymer chains at the lens surface. SFG showed that the lens surface became partially dehydrated upon exposure to air. On this partially dehydrated lens surface, the non-crosslinked domains exhibited low friction and adhesion in AFM. Fully hydrated in saline solution, the non-crosslinked domains extended more than tens of nanometers into solution and were mobile.

  1. Analysis of effect of nanoporous alumina substrate coated with polypyrrole nanowire on cell morphology based on AFM topography.

    Science.gov (United States)

    El-Said, Waleed Ahmed; Yea, Cheol-Heon; Jung, Mi; Kim, Hyuncheol; Choi, Jeong-Woo

    2010-05-01

    In this study, in situ electrochemical synthesis of polypyrrole nanowires with nanoporous alumina template was described. The formation of highly ordered porous alumina substrate was demonstrated with Atomic Force Microscopy (AFM) and Scanning Electron Microscopy (SEM). In addition, Fourier transform infrared analysis confirmed that polypyrrole (PP) nanowires were synthesized by direct electrochemical oxidation of pyrrole. HeLa cancer cells and HMCF normal cells were immobilized on the polypyrrole nanowires/nanoporous alumina substrates to determine the effects of the substrate on the cell morphology, adhesion and proliferation as well as the biocompatibility of the substrate. Cell adhesion and proliferation were characterized using a standard MTT assay. The effects of the polypyrrole nanowires/nanoporous alumina substrate on the cell morphology were studied by AFM. The nanoporous alumina coated with polypyrrole nanowires was found to exhibit better cell adhesion and proliferation than polystyrene petridish, aluminum foil, 1st anodized and uncoated 2nd anodized alumina substrate. This study showed the potential of the polypyrrole nanowires/nanoporous alumina substrate as biocompatibility electroactive polymer substrate for both healthy and cancer cell cultures applications.

  2. A study on the influence of surface active agent molecules on the AFM scanning and imaging of SWCNTs

    Energy Technology Data Exchange (ETDEWEB)

    Xu Ke; Yu Haibo; Tian Xiaojun; Dong Zaili [State Key Laboratory of Robotics, Shenyang Institute of Automation, CAS, Shenyang 110016 (China); Wu Chengdong [Northeastern University, Shenyang 110004 (China)], E-mail: xuke08@sia.cn

    2009-09-01

    The paper proposed a preprocessing method for scanning samples based on rinsing technology, and solved the effective fabrication problem of AFM scanning samples based on facilitated dispersing SWCNTs with SDS surface active agents. First, the ultrasonication oscillation method was applied, and the uniform alignment of SWCNTs in SDS was realized. Then, SDS solution of different concentrations was scanned and imaged with AFM, the influence of SDS solution on the imaging quality of SWCNTs was analyzed, and the method to effectively fabricate scanning samples after dispersing SWCNTs was found. The results of the experiments showed that the preprocessing of the SWCNTs solution scanning samples was the decisive factor to influence SWCNTs imaging quality, that the result of rinsing SWCNTs scanning samples for 5s at 0.1ml/s with di-ionized water was the best, and that with the same rinsing rate and angle, the rinsing di-ionized water quantity would influence the alignment degree of SWCNTs on the substrates and SDS macromolecules' residual quantity on SWCNTs scanning samples.

  3. Physical-mechanical image of the cell surface on the base of AFM data in contact mode

    Science.gov (United States)

    Starodubtseva, M. N.; Starodubtsev, I. E.; Yegorenkov, N. I.; Kuzhel, N. S.; Konstantinova, E. E.; Chizhik, S. A.

    2017-10-01

    Physical and mechanical properties of the cell surface are well-known markers of a cell state. The complex of the parameters characterizing the cell surface properties, such as the elastic modulus (E), the parameters of adhesive (Fa), and friction (Ff) forces can be measured using atomic force microscope (AFM) in a contact mode and form namely the physical-mechanical image of the cell surface that is a fundamental element of the cell mechanical phenotype. The paper aims at forming the physical-mechanical images of the surface of two types of glutaraldehyde-fixed cancerous cells (human epithelial cells of larynx carcinoma, HEp-2c cells, and breast adenocarcinoma, MCF-7 cells) based on the data obtained by AFM in air and revealing the basic difference between them. The average values of friction, elastic and adhesive forces, and the roughness of lateral force maps, as well as dependence of the fractal dimension of lateral force maps on Z-scale factor have been studied. We have revealed that the response of microscale areas of the HEp-2c cell surface having numerous microvilli to external mechanical forces is less expressed and more homogeneous in comparison with the response of MCF-7 cell surface.

  4. AFM Morphology Study of Si1-Y GeY:H Films Deposited by LF PE CVD from Silane-Germane with Different

    Energy Technology Data Exchange (ETDEWEB)

    Sanchez, L; Kosarev, A

    2005-03-28

    The morphology of Si{sub 1-Y} Ge{sub Y}:H films in the range of Y=0.23 to 0.9 has been studied by AFM. The films were deposited by Low Frequency (LF) PE CVD at substrate temperature T{sub s}=300 C and discharge frequency f=110 kHz from silane+germane mixture with and without, Ar and H{sub 2} dilution. The films were deposited on silicon and glass substrates. AFM images were taken and analyzed for 2 x 2 mm{sup 2} area. All the images demonstrated ''grain'' like structure, which was characterized by the height distribution function F(H) average roughness , standard height deviation Rq, lateral correlation length L{sub c} area distribution function F(s), mean grain area , diameter distribution function F(d), and mean grain diameter . The roughness of the films monotonically increases with Y for all dilutions, but more significantly in the films deposited without dilution. L{sub c} continuously grows with Y in the films deposited without dilution, while more complex behavior L{sub c}(Y) is observed in the films deposited with H- or Ar dilution. The sharpness of F(H) characterized by curtosis {gamma} depends on dilution and the sharpest F(H) are for the films deposited with Ar ({gamma}=5.30,Y=0.23) and without dilution ({gamma}=4.3, Y=0.45). Isothermal annealing caused increase of , L{sub c} in the films deposited with H- and Ar dilutions, while in the films prepared without dilution the behavior was more complex, depending on the substrates. Significant narrowing of the height distribution was observed in the films deposited with H dilution or without dilution.

  5. Identification of surface defects and subsurface dopants in a delta-doped system using simultaneous nc-AFM/STM and DFT

    Czech Academy of Sciences Publication Activity Database

    Spadafora, Evan; Berger, Jan; Mutombo, Pingo; Telychko, Mykola; Švec, Martin; Majzik, Zsolt; McLean, A.B.; Jelínek, Pavel

    2014-01-01

    Roč. 118, č. 29 (2014), s. 15744-15753 ISSN 1932-7447 R&D Projects: GA ČR(CZ) GA14-02079S Institutional support: RVO:68378271 Keywords : doping * subsurface defects * nc-AFM * STM * KPFM * DFT Subject RIV: BM - Solid Matter Physics ; Magnetism Impact factor: 4.772, year: 2014

  6. Wetting at the nanometer scale: effects of long-range forces and substrate heterogeneities

    International Nuclear Information System (INIS)

    Checco, Antonio

    2003-01-01

    Wetting phenomena on the nano-scale remain poorly understood in spite of their growing theoretical and practical interest. In this context, the present work aimed at studying partial wetting of nanometer-sized alkane droplets on 'model' surfaces build by self-assembly of organic monolayers. For this purpose a novel technique, based on 'noncontact' Atomic Force Microscopy (AFM), has been developed to image, with minimal artefacts, drops of adjustable size directly condensed on so- lid surfaces. We have thus shown that contact angle of alkanes, wetting a weakly heterogeneous, silanized substrate, noticeably decreases from its macroscopic value for droplets sizes in the submicron range. The line tension, arising in this case from purely dispersive long-range interactions between the liquid and the substrate, is theoretically too weak to be responsible for the observed effect. Therefore we have supposed that contact angle is affected by mesoscopic chemical heterogeneities of the substrate whenever the droplets size becomes sufficiently small. This scenario has been supported by numerical simulations based on a simplified model of the spatial distribution of surface defects. Similar experiments, performed on different substrates (monolayers made of alkane-thiols self-assembled on gold and of alkyl chains covalently bound onto a silicon surface), have also shown that wetting on small scales is strongly affected by minimal physical and chemical surface heterogeneities. Finally, to provide further examples of the potential of the above mentioned AFM technique, we have studied the wettability of nano-structured surfaces and the local wetting properties of hair. (author) [fr

  7. Multiscale patterning of nanocomposite polyelectrolyte/nanoparticle films using inkjet printing and AFM scratching

    International Nuclear Information System (INIS)

    Leigh, S J; Bowen, J; Preece, J A

    2015-01-01

    The fabrication of structured polymer/nanoparticle composite films through a combination of additive, subtractive and self-assembly methodologies is investigated. Consumer grade inkjet printing hardware is employed to deposit cationic polyelectrolytes on (i) hydrophilic and (ii) hydrophobised glass substrates. The hydrophobisation process controls the spreading of the droplets and hence the lateral size of printed features. The printed cationic polyelectrolyte regions are used as a template to direct the self-assembly of negatively charged gold nanoparticles onto the surface. Micro-scale features are created in the polyelectrolyte/nanoparticle films using AFM scratching to selectively displace material. The effect of substrate wettability on film morphology is discussed. (paper)

  8. Study of AFM-based nanometric cutting process using molecular dynamics

    International Nuclear Information System (INIS)

    Zhu Pengzhe; Hu Yuanzhong; Ma Tianbao; Wang Hui

    2010-01-01

    Three-dimensional molecular dynamics (MD) simulations are conducted to investigate the atomic force microscope (AFM)-based nanometric cutting process of copper using diamond tool. The effects of tool geometry, cutting depth, cutting velocity and bulk temperature are studied. It is found that the tool geometry has a significant effect on the cutting resistance. The friction coefficient (cutting resistance) on the nanoscale decreases with the increase of tool angle as predicted by the macroscale theory. However, the friction coefficients on the nanoscale are bigger than those on the macroscale. The simulation results show that a bigger cutting depth results in more material deformation and larger chip volume, thus leading to bigger cutting force and bigger normal force. It is also observed that a higher cutting velocity results in a larger chip volume in front of the tool and bigger cutting force and normal force. The chip volume in front of the tool increases while the cutting force and normal force decrease with the increase of bulk temperature.

  9. XPS, UV–vis spectroscopy and AFM studies on removal mechanisms of Si-face SiC wafer chemical mechanical polishing (CMP)

    International Nuclear Information System (INIS)

    Zhou, Yan; Pan, Guoshun; Shi, Xiaolei; Xu, Li; Zou, Chunli; Gong, Hua; Luo, Guihai

    2014-01-01

    Highlights: • CMP removal mechanism of Si-face SiC wafer is investigated through XPS analysis. • UV–vis spectroscopy is used to study CMP removal mechanisms. • CMP removal model of Si-face SiC wafer is proposed. • The variations of atomic step morphology on ultra-smooth surface via AFM is studied. - Abstract: Chemical mechanical polishing (CMP) removal mechanisms of on-axis Si-face SiC wafer have been investigated through X-ray photoelectron spectroscopy (XPS), UV–visible (UV–vis) spectroscopy and atomic force microscopy (AFM). XPS results indicate that silicon oxide is formed on Si-face surface polished by the slurry including oxidant H 2 O 2 , but not that after immersing in H 2 O 2 solution. UV–vis spectroscopy curves prove that • OH hydroxyl radical could be generated only under CMP polishing by the slurry including H 2 O 2 and abrasive, so as to promote oxidation of Si-face to realize the effective removal; meanwhile, alkali KOH during CMP could induce the production of more radicals to improve the removal. On the other side, ultra-smooth polished surface with atomic step structure morphology and extremely low Ra of about 0.06 nm (through AFM) is obtained using the developed slurry with silica nanoparticle abrasive. Through investigating the variations of the atomic step morphology on the surface polished by different slurries, it's reveals that CMP removal mechanism involves a simultaneous process of surface chemical reaction and nanoparticle atomic scale abrasion

  10. Investigation of geometrical effects in the carbon allotropes manipulation based on AFM: multiscale approach

    Energy Technology Data Exchange (ETDEWEB)

    Korayem, M. H., E-mail: hkorayem@iust.ac.ir; Hefzabad, R. N.; Homayooni, A.; Aslani, H. [Iran University of Science and Technology, Robotic Research Laboratory, Center of Excellence in Experimental Solid Mechanics and Dynamics, School of Mechanical Engineering (Iran, Islamic Republic of)

    2017-01-15

    Carbon allotropes are used as nanocarriers for drug and cell delivery. To obtain an accurate result in the nanoscale, it is important to use a precise model. In this paper, a multiscale approach is presented to investigate the manipulation process of carbon allotropes based on atomic force microscopy (AFM). For this purpose, the AFM setup is separated into two parts with different sizes as macro field (MF) and nano field (NF). Using Kirchhoff’s plate model, the cantilever (the main part of MF) is modeled. The molecular dynamics method is applied to model the NF part, and then the MF and NF are coupled with the multiscale algorithm. With this model, by considering the effect of size and shape, the manipulation of carbon allotropes is carried out. The manipulations of armchair CNTs and fullerenes are performed to study the diameter changing effects. The result shows that the manipulation and friction force increases by increasing the diameter. The result of the indentation depth for the armchair CNTs indicates that decreasing the diameter causes the indentation depth to reduce. Moreover, the manipulations of four kinds of carbon allotropes with the same number of atoms have been studied to investigate the geometrical effects. The shapes of these nanoparticles change from sphere to cylinder. The results illustrate that the manipulation and the friction force decrease as the nanoparticle shape varies from sphere to cylinder. The Von-Mises results demonstrate that by changing the nanoparticle shape from the spherical to the cylindrical form, the stress increases, although the manipulation force reduces.

  11. Investigation of geometrical effects in the carbon allotropes manipulation based on AFM: multiscale approach

    International Nuclear Information System (INIS)

    Korayem, M. H.; Hefzabad, R. N.; Homayooni, A.; Aslani, H.

    2017-01-01

    Carbon allotropes are used as nanocarriers for drug and cell delivery. To obtain an accurate result in the nanoscale, it is important to use a precise model. In this paper, a multiscale approach is presented to investigate the manipulation process of carbon allotropes based on atomic force microscopy (AFM). For this purpose, the AFM setup is separated into two parts with different sizes as macro field (MF) and nano field (NF). Using Kirchhoff’s plate model, the cantilever (the main part of MF) is modeled. The molecular dynamics method is applied to model the NF part, and then the MF and NF are coupled with the multiscale algorithm. With this model, by considering the effect of size and shape, the manipulation of carbon allotropes is carried out. The manipulations of armchair CNTs and fullerenes are performed to study the diameter changing effects. The result shows that the manipulation and friction force increases by increasing the diameter. The result of the indentation depth for the armchair CNTs indicates that decreasing the diameter causes the indentation depth to reduce. Moreover, the manipulations of four kinds of carbon allotropes with the same number of atoms have been studied to investigate the geometrical effects. The shapes of these nanoparticles change from sphere to cylinder. The results illustrate that the manipulation and the friction force decrease as the nanoparticle shape varies from sphere to cylinder. The Von-Mises results demonstrate that by changing the nanoparticle shape from the spherical to the cylindrical form, the stress increases, although the manipulation force reduces.

  12. AFM study of surface relief evolution in 316L steel fatigued at low and high temperatures

    Czech Academy of Sciences Publication Activity Database

    Man, Jiří; Valtr, B.; Weidner, A.; Petrenec, Martin; Obrtlík, Karel; Polák, Jaroslav

    2010-01-01

    Roč. 2, č. 1 (2010), s. 1625-1633 E-ISSN 1877-7058. [Fatigue 2010. Praha, 06.06.2010-11.06.2010] R&D Projects: GA ČR GAP108/10/2371; GA AV ČR 1QS200410502; GA ČR GA106/06/1096 Institutional research plan: CEZ:AV0Z20410507 Keywords : Fatigue crack initiation * 316L steel * Persistent slip band (PSB) * Extrusion * Intrusion * Atomic force microscopy (AFM) Subject RIV: JL - Materials Fatigue, Friction Mechanics

  13. Chemical Etching, AFM, Laser Damage Threshold, and Nonlinear Optical Studies of Potential Nonlinear Optical Crystal: Bis (L-Glutamine Potassium Nitrate

    Directory of Open Access Journals (Sweden)

    Redrothu Hanumantharao

    2013-01-01

    Full Text Available A novel semiorganic nonlinear optical crystal bis (L-glutamine potassium nitrate (BGPN grown by slow evaporation technique at ambient temperature. The grown crystal surface has been analyzed by chemical etching and atomic force microscopy (AFM studies. Amplitude parameters like area roughness, roughness average, valley height, valley depth, peak height, and peak valley height were measured successfully from AFM studies. Etching studies were carried out by various solvents like water, methanol and ethanol. The etching study indicates the occurrence of different types of etch pit patterns like striations and steplike pattern. The laser damage threshold energy has been measured by irradiating laser beam using a Q-switched Nd: YAG laser (1064 nm. Second harmonic generation (SHG studies have been performed by famous Kurtz powder technique with reference to standard potassium dihydrogen phosphate single crystals (KDP. It is found from this technique that SHG efficiency of BGPN is in comparison to that of standard KDP crystals.

  14. Continuous AFM observation of martensitic transformation and its reversion in training cycles of Fe-Mn-Si based shape memory alloys

    International Nuclear Information System (INIS)

    Liu, D.Z.; Kikuchi, T.; Kajiwara, S.; Shinya, N.

    2000-01-01

    The effect of thermomechanical treatment (so-called ''training'') cycles on stress-induced martensitic transformation and its reversion has been studied by atomic force microscopy (AFM) to make clear the origin of improvement of shape memory effect (SME) due to training in Fe-Mn-Si based shape memory alloys (SMAs). It was found that training cycles make martensite plates tend to grow on the primary shear system, i.e., on the most favorable shear system for the fcc to hcp martensitic transformation. In addition, training cycles lead to a more uniform distribution of thin martensite plates in a grain. Martensitic plates with the above characteristics are easier to be reverted back to parent phase when heated, and then nearly perfect SME is obtained. AFM observation shows that the key factor to realize perfect SME in Fe-Mn-Si based SMAs is to produce the uniform distribution of thin martensite plates on the primary shear system when deformed by external stress. (orig.)

  15. Aggregation mechanism of Pd nanoparticles in L-cysteine aqueous solution studied by NEXAFS and AFM

    International Nuclear Information System (INIS)

    Tsukada, C.; Ogawa, S.; Mizutani, T.; Kutluk, G.; Namatame, H.; Taniguchi, M.; Yagi, S.

    2012-01-01

    Highlight: ► We focus on the biocompatibility of Pd nanoparticles (NPs) for L-cysteine under water environment. ► The Pd NPs have been fabricated and deposited on Si wafer by gas evaporation method. ► When the Pd NPs/Si has been dipped into L-cysteine aqueous solution, the L-cysteine has selectively adsorbed on Pd NPs surface and existed as the L-cysteine thiolate, atomic S and L-cystine. ► Moreover, the aggregation of Pd NPs occurs by the migration of Pd NPs on Si and the cross-linked reaction between L-cysteine thiolate molecules adsorbed on Pd NPs. - Abstract: We focus on the biocompatibility of Pd nanoparticles (NPs) from the point of microscopic view. Thus, as the basic research for the biocompatibility, we have investigated the adsorbates on the Pd NPs surface and the aggregation mechanism for the Pd NPs on Si substrate after dipping into L-cysteine aqueous solution by means of NEXAFS measurement and AFM observation. The Pd NPs have been fabricated and deposited on the Si wafer by the gas evaporation method. Judging from the results of NEXAFS measurement, it is clear that the L-cysteine thiolate and atomic S exist on the Pd NPs surface. The results of AFM observation show that the Pd NPs aggregate. It is thought that the aggregation of the Pd NPs occurs by both the migration of the Pd NPs on Si wafer and the cross-linked reaction.

  16. Simultaneous Measurement of Multiple Mechanical Properties of Single Cells Using AFM by Indentation and Vibration.

    Science.gov (United States)

    Zhang, Chuang; Shi, Jialin; Wang, Wenxue; Xi, Ning; Wang, Yuechao; Liu, Lianqing

    2017-12-01

    The mechanical properties of cells, which are the main characteristics determining their physical performance and physiological functions, have been actively studied in the fields of cytobiology and biomedical engineering and for the development of medicines. In this study, an indentation-vibration-based method is proposed to simultaneously measure the mechanical properties of cells in situ, including cellular mass (m), elasticity (k), and viscosity (c). The proposed measurement method is implemented based on the principle of forced vibration stimulated by simple harmonic force using an atomic force microscope (AFM) system integrated with a piezoelectric transducer as the substrate vibrator. The corresponding theoretical model containing the three mechanical properties is derived and used to perform simulations and calculations. Living and fixed human embryonic kidney 293 (HEK 293) cells were subjected to indentation and vibration to measure and compare their mechanical parameters and verify the proposed approach. The results that the fixed sample cells are more viscous and elastic than the living sample cells and the measured mechanical properties of cell are consistent within, but not outside of the central region of the cell, are in accordance with the previous studies. This work provides an approach to simultaneous measurement of the multiple mechanical properties of single cells using an integrated AFM system based on the principle force vibration and thickness-corrected Hertz model. This study should contribute to progress in biomedical engineering, cytobiology, medicine, early diagnosis, specific therapy and cell-powered robots.

  17. Calibration of a large multi-element neutron counter in the energy range 85-430 MeV

    CERN Document Server

    Strong, J A; Esterling, R J; Garvey, J; Green, M G; Harnew, N; Jane, M R; Jobes, M; Mawson, J; McMahon, T; Robertson, A W; Thomas, D H

    1978-01-01

    Describes the calibration of a large 60 element neutron counter with a threshold of 2.7 MeV equivalent electron energy. The performance of the counter has been measured in the neutron kinetic energy range 8.5-430 MeV using a neutron beam at the CERN Synchrocyclotron. The results obtained for the efficiency as a function of energy are in reasonable agreement with a Monte Carlo calculation. (7 refs).

  18. An objective approach to determining the weight ranges of prey preferred by and accessible to the five large African carnivores.

    Directory of Open Access Journals (Sweden)

    Hayley S Clements

    Full Text Available Broad-scale models describing predator prey preferences serve as useful departure points for understanding predator-prey interactions at finer scales. Previous analyses used a subjective approach to identify prey weight preferences of the five large African carnivores, hence their accuracy is questionable. This study uses a segmented model of prey weight versus prey preference to objectively quantify the prey weight preferences of the five large African carnivores. Based on simulations of known predator prey preference, for prey species sample sizes above 32 the segmented model approach detects up to four known changes in prey weight preference (represented by model break-points with high rates of detection (75% to 100% of simulations, depending on number of break-points and accuracy (within 1.3±4.0 to 2.7±4.4 of known break-point. When applied to the five large African carnivores, using carnivore diet information from across Africa, the model detected weight ranges of prey that are preferred, killed relative to their abundance, and avoided by each carnivore. Prey in the weight ranges preferred and killed relative to their abundance are together termed "accessible prey". Accessible prey weight ranges were found to be 14-135 kg for cheetah Acinonyx jubatus, 1-45 kg for leopard Panthera pardus, 32-632 kg for lion Panthera leo, 15-1600 kg for spotted hyaena Crocuta crocuta and 10-289 kg for wild dog Lycaon pictus. An assessment of carnivore diets throughout Africa found these accessible prey weight ranges include 88±2% (cheetah, 82±3% (leopard, 81±2% (lion, 97±2% (spotted hyaena and 96±2% (wild dog of kills. These descriptions of prey weight preferences therefore contribute to our understanding of the diet spectrum of the five large African carnivores. Where datasets meet the minimum sample size requirements, the segmented model approach provides a means of determining, and comparing, the prey weight range preferences of any carnivore

  19. An objective approach to determining the weight ranges of prey preferred by and accessible to the five large African carnivores.

    Science.gov (United States)

    Clements, Hayley S; Tambling, Craig J; Hayward, Matt W; Kerley, Graham I H

    2014-01-01

    Broad-scale models describing predator prey preferences serve as useful departure points for understanding predator-prey interactions at finer scales. Previous analyses used a subjective approach to identify prey weight preferences of the five large African carnivores, hence their accuracy is questionable. This study uses a segmented model of prey weight versus prey preference to objectively quantify the prey weight preferences of the five large African carnivores. Based on simulations of known predator prey preference, for prey species sample sizes above 32 the segmented model approach detects up to four known changes in prey weight preference (represented by model break-points) with high rates of detection (75% to 100% of simulations, depending on number of break-points) and accuracy (within 1.3±4.0 to 2.7±4.4 of known break-point). When applied to the five large African carnivores, using carnivore diet information from across Africa, the model detected weight ranges of prey that are preferred, killed relative to their abundance, and avoided by each carnivore. Prey in the weight ranges preferred and killed relative to their abundance are together termed "accessible prey". Accessible prey weight ranges were found to be 14-135 kg for cheetah Acinonyx jubatus, 1-45 kg for leopard Panthera pardus, 32-632 kg for lion Panthera leo, 15-1600 kg for spotted hyaena Crocuta crocuta and 10-289 kg for wild dog Lycaon pictus. An assessment of carnivore diets throughout Africa found these accessible prey weight ranges include 88±2% (cheetah), 82±3% (leopard), 81±2% (lion), 97±2% (spotted hyaena) and 96±2% (wild dog) of kills. These descriptions of prey weight preferences therefore contribute to our understanding of the diet spectrum of the five large African carnivores. Where datasets meet the minimum sample size requirements, the segmented model approach provides a means of determining, and comparing, the prey weight range preferences of any carnivore species.

  20. AFM Imaging of Hybridization Chain Reaction-Mediated Signal Transmission Between two DNA Origami Structures

    DEFF Research Database (Denmark)

    Helmig, Sarah Wendelbo; Gothelf, Kurt Vesterager

    2017-01-01

    transfer between two connected DNA nanostructures, using the hybridization chain reaction (HCR). Two sets of metastable DNA hairpins - of which one is immobilized in specific points along tracks on DNA origami structures - are polymerized to form a continuous DNA duplex, which is visible using atomic force...... microscopy (AFM). Upon addition of a designed initiator, the initiation signal is efficiently transferred >200 nm from a specific location on one origami structure to an end point on another origami structure. The system shows no significant loss of signal when crossing from one nanostructure to another...

  1. A high resolution large dynamic range TDC circuit implementation

    International Nuclear Information System (INIS)

    Lei Wuhu; Liu Songqiu; Ye Weiguo; Han Hui; Li Pengyu

    2003-01-01

    Time measurement technology is usually used in nuclear experimentation. There are many methods of time measurement. The implementation method of Time to Digital Conversion (TDC) by means of electronic is a classical technology. The range and resolution of TDC is different according with different usage. A wide range and high resolution TDC circuit, including its theory and implementation way, is introduced in this paper. The test result is also given. (authors)

  2. A high resolution large dynamic range TDC circuit implementation

    International Nuclear Information System (INIS)

    Lei Wuhu; Liu Songqiu; Li Pengyu; Han Hui; Ye Yanlin

    2005-01-01

    Time measurement technology is usually used in nuclear experimentation. There are many methods of time measurement. The implementation method of Time to Digital Conversion (TDC) by means of electronics is a classical technology. The range and resolution of TDC is different according with different usage. A wide range and high resolution TDC circuit, including its theory and implementation way, is introduced in this paper. The test result is also given. (authors)

  3. Quantitative multichannel NC-AFM data analysis of graphene growth on SiC(0001

    Directory of Open Access Journals (Sweden)

    Christian Held

    2012-02-01

    Full Text Available Noncontact atomic force microscopy provides access to several complementary signals, such as topography, damping, and contact potential. The traditional presentation of such data sets in adjacent figures or in colour-coded pseudo-three-dimensional plots gives only a qualitative impression. We introduce two-dimensional histograms for the representation of multichannel NC-AFM data sets in a quantitative fashion. Presentation and analysis are exemplified for topography and contact-potential data for graphene grown epitaxially on 6H-SiC(0001, as recorded by Kelvin probe force microscopy in ultrahigh vacuum. Sample preparations by thermal decomposition in ultrahigh vacuum and in an argon atmosphere are compared and the respective growth mechanisms discussed.

  4. Molecular shape and binding force of Mycoplasma mobile's leg protein Gli349 revealed by an AFM study

    International Nuclear Information System (INIS)

    Lesoil, Charles; Nonaka, Takahiro; Sekiguchi, Hiroshi; Osada, Toshiya; Miyata, Makoto; Afrin, Rehana; Ikai, Atsushi

    2010-01-01

    Recent studies of the gliding bacteria Mycoplasma mobile have identified a family of proteins called the Gli family which was considered to be involved in this novel and yet fairly unknown motility system. The 349 kDa protein called Gli349 was successfully isolated and purified from the bacteria, and electron microscopy imaging and antibody experiments led to the hypothesis that it acts as the 'leg' of M. mobile, responsible for attachment to the substrate as well as for gliding motility. However, more precise evidence of the molecular shape and function of this protein was required to asses this theory any further. In this study, an atomic force microscope (AFM) was used both as an imaging and a force measurement device to provide new information about Gli349 and its role in gliding motility. AFM images of the protein were obtained revealing a complex structure with both rigid and flexible parts, consistent with previous electron micrographs of the protein. Single-molecular force spectroscopy experiments were also performed, revealing that Gli349 is able to specifically bind to sialyllactose molecules and withstand unbinding forces around 70 pN. These findings strongly support the idea that Gli349 is the 'leg' protein of M. mobile, responsible for binding and also most probably force generation during gliding motility.

  5. AFM-based detection of glycocalyx degradation and endothelial stiffening in the db/db mouse model of diabetes.

    Science.gov (United States)

    Targosz-Korecka, Marta; Jaglarz, Magdalena; Malek-Zietek, Katarzyna E; Gregorius, Aleksandra; Zakrzewska, Agnieszka; Sitek, Barbara; Rajfur, Zenon; Chlopicki, Stefan; Szymonski, Marek

    2017-11-21

    Degradation of the glycocalyx and stiffening of endothelium are important pathophysiological components of endothelial dysfunction. However, to our knowledge, these events have not been investigated in tandem in experimental diabetes. Here, the mechanical properties of the glycocalyx and endothelium in ex vivo mouse aorta were determined simultaneously in indentation experiments with an atomic force microscope (AFM) for diabetic db/db and control db/+ mice at ages of 11-19 weeks. To analyze highly heterogeneous aorta samples, we developed a tailored classification procedure of indentation data based on a bi-layer brush model supplemented with Hertz model for quantification of nanomechanics of endothelial regions with and without the glycocalyx surface. In db/db mice, marked endothelial stiffening and reduced glycocalyx coverage were present already in 11-week-old mice and persisted in older animals. In contrast, reduction of the effective glycocalyx length was progressive and was most pronounced in 19-week-old db/db mice. The reduction of the glycocalyx length correlated with an increasing level of glycated haemoglobin and decreased endothelial NO production. In conclusion, AFM nanoindentation analysis revealed that stiffening of endothelial cells and diminished glycocalyx coverage occurred in early diabetes and were followed by the reduction of the glycocalyx length that correlated with diabetes progression.

  6. Preamplifier development for high count-rate, large dynamic range readout of inorganic scintillators

    Energy Technology Data Exchange (ETDEWEB)

    Keshelashvili, Irakli; Erni, Werner; Steinacher, Michael; Krusche, Bernd; Collaboration: PANDA-Collaboration

    2013-07-01

    Electromagnetic calorimeter are central component of many experiments in nuclear and particle physics. Modern ''trigger less'' detectors run with very high count-rates, require good time and energy resolution, and large dynamic range. In addition photosensors and preamplifiers must work in hostile environments (magnetic fields). Due to later constraints mainly Avalanche Photo Diodes (APD's), Vacuum Photo Triodes (VPT's), and Vacuum Photo Tetrodes (VPTT's) are used. A disadvantage is their low gain which together with other requirements is a challenge for the preamplifier design. Our group has developed special Low Noise / Low Power (LNP) preamplifier for this purpose. They will be used to equip PANDA EMC forward end-cap (dynamic range 15'000, rate 1MHz), where the PWO II crystals and preamplifier have to run in an environment cooled down to -25{sup o}C. Further application is the upgrade of the Crystal Barrel detector at the Bonn ELSA accelerator with APD readout for which special temperature comparison of the APD gain and good time resolution is necessary. Development and all test procedures after the mass production done by our group during past several years in Basel University will be reported.

  7. AFM study of adsorption of protein A on a poly(dimethylsiloxane) surface

    International Nuclear Information System (INIS)

    Yu Ling; Lu Zhisong; Gan Ye; Liu Yingshuai; Li, C M

    2009-01-01

    In this paper, the morphology and kinetics of adsorption of protein A on a PDMS surface is studied by AFM. The results of effects of pH, protein concentration and contact time of the adsorption reveal that the morphology of adsorbed protein A is significantly affected by pH and adsorbed surface concentration, in which the pH away from the isoelectric point (IEP) of protein A could produce electrical repulsion to change the protein conformation, while the high adsorbed surface protein volume results in molecular networks. Protein A can form an adsorbed protein film on PDMS with a maximum volume of 2.45 x 10 -3 μm 3 . This work enhances our fundamental understanding of protein A adsorption on PDMS, a frequently used substrate component in miniaturized immunoassay devices.

  8. Nanospot soldering polystyrene nanoparticles with an optical fiber probe laser irradiating a metallic AFM probe based on the near-field enhancement effect.

    Science.gov (United States)

    Cui, Jianlei; Yang, Lijun; Wang, Yang; Mei, Xuesong; Wang, Wenjun; Hou, Chaojian

    2015-02-04

    With the development of nanoscience and nanotechnology for the bottom-up nanofabrication of nanostructures formed from polystyrene nanoparticles, joining technology is an essential step in the manufacturing and assembly of nanodevices and nanostructures in order to provide mechanical integration and connection. To study the nanospot welding of polystyrene nanoparticles, we propose a new nanospot-soldering method using the near-field enhancement effect of a metallic atomic force microscope (AFM) probe tip that is irradiated by an optical fiber probe laser. On the basis of our theoretical analysis of the near-field enhancement effect, we set up an experimental system for nanospot soldering; this approach is carried out by using an optical fiber probe laser to irradiate the AFM probe tip to sinter the nanoparticles, providing a promising technical approach for the application of nanosoldering in nanoscience and nanotechnology.

  9. Periodic long-range transport in a large volume dc glow discharge dusty plasma

    International Nuclear Information System (INIS)

    Thomas, Edward Jr.; Amatucci, William E.; Compton, Christopher; Christy, Brian; Jackson, Jon David

    2003-01-01

    In an earlier paper, the authors reported on observations of a variety of particle transport phenomena observed in DUPLEX--the DUsty PLasma EXperiment at the Naval Research Laboratory [E. Thomas, Jr., W. E. Amatucci, C. Compton, and B. Christy, Phys. Plasmas 9, 3154 (2002)]. DUPLEX is a large, transparent polycarbonate cylinder that is 40 cm in radius and 80 cm in height. dc glow discharge argon plasmas are generated in DUPLEX. In this paper, the authors expand upon one particular feature of particle transport in DUPLEX, the long-range (i.e., greater than 15 cm), periodic (T∼2.5 min) transport of suspended alumina particles through the plasma. A detailed description of this particle motion through the plasma is presented. Finally, a qualitative model describing the phenomena that lead to this transport is also given

  10. Association between Urinary Aflatoxin (AFM1) and Dietary Intake among Adults in Hulu Langat District, Selangor, Malaysia

    OpenAIRE

    Siti Husna Sulaiman; Rosita Jamaluddin; Mohd Redzwan Sabran

    2018-01-01

    Aflatoxin is a food contaminant and its exposure through the diet is frequent and ubiquitous. A long-term dietary aflatoxin exposure has been linked to the development of liver cancer in populations with high prevalence of aflatoxin contamination in foods. Therefore, this study was conducted to identify the association between urinary aflatoxin M1 (AFM1), a biomarker of aflatoxin exposure, with the dietary intake among adults in Hulu Langat district, Selangor, Malaysia. Certain food products ...

  11. Nanomechanical characterization of nanostructured bainitic steel: Peak Force Microscopy and Nanoindentation with AFM.

    Science.gov (United States)

    Morales-Rivas, Lucia; González-Orive, Alejandro; Garcia-Mateo, Carlos; Hernández-Creus, Alberto; Caballero, Francisca G; Vázquez, Luis

    2015-11-25

    The full understanding of the deformation mechanisms in nanostructured bainite requires the local characterization of its mechanical properties, which are expected to change from one phase, bainitic ferrite, to another, austenite. This study becomes a challenging process due to the bainitic nanostructured nature and high Young's modulus. In this work, we have carried out such study by means of the combination of AFM-based techniques, such as nanoindentation and Peak Force Quantitative Nanomechanical Mapping (PF-QNM) measurements. We have addressed critically the limits and advantages of these techniques and been able to measure some elastoplastic parameters of both phases. Specifically, we have analyzed by PF-QNM two nanostructured bainitic steels, with a finer and a coarser structure, and found that both phases have a similar Young's modulus.

  12. Quenched Large Deviations for Simple Random Walks on Percolation Clusters Including Long-Range Correlations

    Science.gov (United States)

    Berger, Noam; Mukherjee, Chiranjib; Okamura, Kazuki

    2018-03-01

    We prove a quenched large deviation principle (LDP) for a simple random walk on a supercritical percolation cluster (SRWPC) on {Z^d} ({d ≥ 2}). The models under interest include classical Bernoulli bond and site percolation as well as models that exhibit long range correlations, like the random cluster model, the random interlacement and the vacant set of random interlacements (for {d ≥ 3}) and the level sets of the Gaussian free field ({d≥ 3}). Inspired by the methods developed by Kosygina et al. (Commun Pure Appl Math 59:1489-1521, 2006) for proving quenched LDP for elliptic diffusions with a random drift, and by Yilmaz (Commun Pure Appl Math 62(8):1033-1075, 2009) and Rosenbluth (Quenched large deviations for multidimensional random walks in a random environment: a variational formula. Ph.D. thesis, NYU, arXiv:0804.1444v1) for similar results regarding elliptic random walks in random environment, we take the point of view of the moving particle and prove a large deviation principle for the quenched distribution of the pair empirical measures of the environment Markov chain in the non-elliptic case of SRWPC. Via a contraction principle, this reduces easily to a quenched LDP for the distribution of the mean velocity of the random walk and both rate functions admit explicit variational formulas. The main difficulty in our set up lies in the inherent non-ellipticity as well as the lack of translation-invariance stemming from conditioning on the fact that the origin belongs to the infinite cluster. We develop a unifying approach for proving quenched large deviations for SRWPC based on exploiting coercivity properties of the relative entropies in the context of convex variational analysis, combined with input from ergodic theory and invoking geometric properties of the supercritical percolation cluster.

  13. A Doherty Power Amplifier with Large Back-Off Power Range Using Integrated Enhancing Reactance

    Directory of Open Access Journals (Sweden)

    Wa Kong

    2018-01-01

    Full Text Available A symmetric Doherty power amplifier (DPA based on integrated enhancing reactance (IER was proposed for large back-off applications. The IER was generated using the peaking amplifier with the help of a desired impedance transformation in the low-power region to enhance the back-off efficiency of the carrier amplifier. To convert the impedances properly, both in the low-power region and at saturation, a two-impedance matching method was employed to design the output matching networks. For verification, a symmetric DPA with large back-off power range over 2.2–2.5 GHz was designed and fabricated. Measurement results show that the designed DPA has the 9 dB back-off efficiency of higher than 45%, while the saturated output power is higher than 44 dBm over the whole operation bandwidth. When driven by a 20 MHz LTE signal, the DPA can achieve good average efficiency of around 50% with adjacent channel leakage ratio of about –50 dBc after linearization over the frequency band of interest. The linearity improvement of the DPA for multistandard wireless communication system was also verified with a dual-band modulated signal.

  14. Acquire an Bruker Dimension FastScanTM Atomic Force Microscope (AFM) for Materials, Physical and Biological Science Research and Education

    Science.gov (United States)

    2016-04-14

    study dynamic events such as melting, evaporation, crystallization, dissolution, self-assembly, membrane disruption, sample movement tracking. To... polymeric hairy nanopraticle, suprastructures REPORT DOCUMENTATION PAGE 11. SPONSOR/MONITOR’S REPORT NUMBER(S) 10. SPONSOR/MONITOR’S ACRONYM(S...the AFM will permit us to study dynamic events such as melting, evaporation, crystallization, dissolution, self-assembly, membrane disruption, sample

  15. Large magnetocurrents in double-barrier tunneling transistors

    International Nuclear Information System (INIS)

    Lee, J.H.; Jun, K.-I.; Shin, K.-H.; Park, S.Y.; Hong, J.K.; Rhie, K.; Lee, B.C.

    2005-01-01

    Magnetic tunneling transistors (MTT) with double tunneling barriers are fabricated. The structure of the transistor is AFM/FM/I/FM/I/FM/AFM, and ferromagnetic layers serve as the emitter, base and collector. This double-barrier tunneling transistor (DBTT) has an advantage of controlling the potential between the base and collector, compared to the Schottky-barrier-based base and collector of MTT. We found that the collector current density of DBTT is at least 10 3 times larger than that of conventional MTT, since tunneling through AlO x barrier provides much larger current density than that through Schottky barrier

  16. Surface analysis of glass fibres using XPS and AFM: case study of glass fibres recovered from the glass fibre reinforced polymer using chemical recycling

    Science.gov (United States)

    Nzioka, A. M.; Kim, Y. J.

    2018-01-01

    In this study, we present the results of an experimental study of the use of the X-ray photoelectron spectroscopy (XPS) and atomic force microscopy (AFM) to characterise the coatings of the recovered E - glass fibres. The recovered E - glass fibres were obtained using chemical recycling process coupled with ultrasound cavitation. The objective of this study was to analyse the impact of chemical recycling and the ultrasound cavitation process on the sizing properties of the recovered fibres. We obtained the recovered fibres and sized using 1 wt% 3 - aminopropyltriethoxysilane (APS). Part of the sized fibres was washed with acetone and analysed all the sample fibres using AFM and XPS. Results showed the different composition of sizing after extraction using acetone. We compared the results of this study with that of virgin clean glass fibres.

  17. Optical imaging beyond the diffraction limit by SNEM: Effects of AFM tip modifications with thiol monolayers on imaging quality

    Energy Technology Data Exchange (ETDEWEB)

    Cumurcu, Aysegul [Materials Science and Technology of Polymers, MESA+ Institute for Nanotechnology, University of Twente, Enschede NL-7500 (Netherlands); Dutch Polymer Institute (DPI), P.O. Box 902, 5600 AX, Eindhoven (Netherlands); Diaz, Jordi [Scientific and Technological Centers of the University of Barcelona, C/ Lluís Solé i Sabaris, 1-3, 08028 Barcelona (Spain); Lindsay, Ian D. [Nanophysics and Soft Matter Group, H.H. Wills Physics Laboratory, University of Bristol, Tyndall Avenue, Bristol BS8 1TL (United Kingdom); Beer, Sissi de; Duvigneau, Joost [Materials Science and Technology of Polymers, MESA+ Institute for Nanotechnology, University of Twente, Enschede NL-7500 (Netherlands); Schön, Peter [Materials Science and Technology of Polymers, MESA+ Institute for Nanotechnology, University of Twente, Enschede NL-7500 (Netherlands); NanoBioInterface, Research Center Design and Technology, Saxion University of Applied Sciences, 7500 KB Enschede (Netherlands); Julius Vancso, G., E-mail: g.j.vancso@utwente.nl [Materials Science and Technology of Polymers, MESA+ Institute for Nanotechnology, University of Twente, Enschede NL-7500 (Netherlands)

    2015-03-15

    Tip-enhanced nanoscale optical imaging techniques such as apertureless scanning near-field optical microscopy (a-SNOM) and scanning near-field ellipsometric microscopy (SNEM) applications can suffer from a steady degradation in performance due to adhesion of atmospheric contaminants to the metal coated tip. Here, we demonstrate that a self-assembled monolayer (SAM) of ethanethiol (EtSH) is an effective means of protecting gold-coated atomic force microscopy (AFM) probe tips from accumulation of surface contaminants during prolonged exposure to ambient air. The period over which they yield consistent and reproducible results for scanning near-field ellipsometric microscopy (SNEM) imaging is thus extended. SNEM optical images of a microphase separated polystyrene-block-poly (methylmethacrylate) (PS-b-PMMA) diblock copolymer film, which were captured with bare and SAM-protected gold-coated AFM probes, both immediately after coating and following five days of storage in ambient air, were compared. During this period the intensity of the optical signals from the untreated gold tip fell by 66%, while those from the SAM protected tip fell by 14%. Additionally, gold coated AFM probe tips were modified with various lengths of alkanethiols to measure the change in intensity variation in the optical images with SAM layer thickness. The experimental results were compared to point dipole model calculations. While a SAM of 1-dodecanethiol (DoSH) was found to strongly suppress field enhancement we find that it can be locally removed from the tip apex by deforming the molecules under load, restoring SNEM image contrast. - Highlights: • SAM of ethanethiol is used to prevent contamination of gold coated tips. • Functionalizing gold coated tips with a SAM lead to reproducible SNEM imaging. • Point dipole model agreed with the experimental results of the SNEM images. • SAM of 1-dodecanethiol was found to strongly suppress field enhancement in SNEM. • SAM of 1-dodecanethiol

  18. Microelectromechanical system device for calibration of atomic force microscope cantilever spring constants between 0.01 and 4 N/m

    International Nuclear Information System (INIS)

    Cumpson, Peter J.; Hedley, John; Clifford, Charles A.; Chen Xinyong; Allen, Stephanie

    2004-01-01

    Calibration of atomic force microscope (AFM) cantilevers is necessary for the measurement of nano-newton and pico-newton forces, which are critical to analytical application of AFM in the analysis of polymer surfaces, biological structures and organic molecules. Previously we have described microfabricated array of reference spring (MARS) devices for AFM cantilever spring-constant calibration. Hitherto, these have been limited to the calibration of AFM cantilevers above 0.03 N/m, although they can be used to calibrate cantilevers of lower stiffness with reduced accuracy. Below this limit MARS devices similar to the designs hitherto described would be fragile and difficult to manufacture with reasonable yield. In this work we describe a device we call torsional MARS. This is a large-area torsional mechanical resonator, manufactured by bulk micromachining of a 'silicon-on-insulator' wafer. By measuring its torsional oscillation accurately in vacuum we can deduce its torsional spring constant. The torsional reference spring spans the range of spring constant (from 4 down to 0.01 N/m) that is important in biological AFM, allowing even the most compliant AFM cantilever to be calibrated easily and rapidly

  19. How do low dispersal species establish large range sizes? The case of the water beetle Graphoderus bilineatus

    DEFF Research Database (Denmark)

    Iversen, Lars Lønsmann; Rannap, Riinu; Thomsen, Philip Francis

    2013-01-01

    important than species phylogeny or local spatial attributes. In this study we used the water beetle Graphoderus bilineatus a philopatric species of conservation concern in Europe as a model to explain large range size and to support effective conservation measures for such species that also have limited...... systems and wetlands which used to be highly connected throughout the central plains of Europe. Our data suggest that a broad habitat niche can prevent landscape elements from becoming barriers for species like G. bilineatus. Therefore, we question the usefulness of site protection as conservation...... measures for G. bilineatus and similar philopatric species. Instead, conservation actions should be focused at the landscape level to ensure a long-term viability of such species across their range....

  20. Corals through the light : phylogenetics, functional diversity and adaptive strategies of coral-symbiont associations over a large depth range

    NARCIS (Netherlands)

    Rodrigues Frade, P.

    2009-01-01

    Light constitutes the main energy source in the coral reef ecosystem, with its intensity dramatically reduced with increasing depth over the reef slope. How do corals thrive across these acute light gradients that exist over large depth ranges? This and many more questions are addressed throughout

  1. AFM visualization of sub-50nm polyplex disposition to the nuclear pore complex without compromising the integrity of the nuclear envelope

    DEFF Research Database (Denmark)

    Andersen, Helene; Parhamifar, Ladan; Hunter, A Christy

    2016-01-01

    that were microinjected into the oocytes of Xenopus laevis, as an example of a non-dividing cell, is exclusive to the nuclear pore complex (NPC). AFM images show NPCs clogged only with sub-50nm polyplexes. This mode of disposition neither altered the morphology/integrity of the nuclear membrane nor the NPC...

  2. Use of a cocktail of spin traps for fingerprinting large range of free radicals in biological systems.

    Science.gov (United States)

    Marchand, Valérie; Charlier, Nicolas; Verrax, Julien; Buc-Calderon, Pedro; Levêque, Philippe; Gallez, Bernard

    2017-01-01

    It is well established that the formation of radical species centered on various atoms is involved in the mechanism leading to the development of several diseases or to the appearance of deleterious effects of toxic molecules. The detection of free radical is possible using Electron Paramagnetic Resonance (EPR) spectroscopy and the spin trapping technique. The classical EPR spin-trapping technique can be considered as a "hypothesis-driven" approach because it requires an a priori assumption regarding the nature of the free radical in order to select the most appropriate spin-trap. We here describe a "data-driven" approach using EPR and a cocktail of spin-traps. The rationale for using this cocktail was that it would cover a wide range of biologically relevant free radicals and have a large range of hydrophilicity and lipophilicity in order to trap free radicals produced in different cellular compartments. As a proof-of-concept, we validated the ability of the system to measure a large variety of free radicals (O-, N-, C-, or S- centered) in well characterized conditions, and we illustrated the ability of the technique to unambiguously detect free radical production in cells exposed to chemicals known to be radical-mediated toxic agents.

  3. Corrosion behaviour of Ni in aprotic solvents an electrochemical, XPS and AFM study

    International Nuclear Information System (INIS)

    Bellucci, F.; Monetta, T.; Capobianco, G.; Deganello, A.; Glisenti, A.; Moretti, G.

    1998-01-01

    Electrochemical and X-ray photoelectron spectroscopic (XPS) techniques have been used to study the passivation of nickel in 0.1 M H 2 SO 4 DMF and ACN solutions with different water content. Electrochemical results indicate the anodic formation of a thin, poor protective layer and the possibility of salt precipitation onto the metallic surface. ARXPS results indicate that while in the anodic film formed in DMF, Ni(OH) 2 constitute the superficial component under which a discontinuous layer of NiO and NiSO 4 is present. Ni(OH) 2 and NiSO 4 are the more superficial constituents in the passivation layer formed in ACN, while NiO becomes prevalent in the underlying layers. AFM images show that in both the solvents the sample surface is very dishomogeneous with flakes and fractures. (orig.)

  4. AFM study of the morphologic change of HDPE surface photografted with glycidyl methacrylate.

    Science.gov (United States)

    Wang, Huiliang; Han, Jianmei

    2009-05-01

    The UV-induced grafting of glycidyl methacrylate (GMA) onto high-density polyethylene (HDPE) and the atomic force microscopy (AFM) study of the morphologic change of the grafted surface are reported. The grafting was carried out in GMA acetone solutions with different monomer concentrations. Grafting was much faster in a solution with a higher monomer concentration. FTIR analyses proved that GMA had been successfully grafted onto HDPE. The morphologies of grafted HDPE surfaces changed with UV irradiation time. The monomer concentration had a significant effect on the morphologies of the grafted HDPE surfaces. The HDPE surface grafted in a solution with a higher monomer concentration was much rougher than that grafted in a solution with a lower monomer concentration. The growth models of the grafted granules or clusters are also proposed.

  5. A control approach to cross-coupling compensation of piezotube scanners in tapping-mode atomic force microscope imaging.

    Science.gov (United States)

    Wu, Ying; Shi, Jian; Su, Chanmin; Zou, Qingze

    2009-04-01

    In this article, an approach based on the recently developed inversion-based iterative control (IIC) to cancel the cross-axis coupling effect of piezoelectric tube scanners (piezoscanners) in tapping-mode atomic force microscope (AFM) imaging is proposed. Cross-axis coupling effect generally exists in piezoscanners used for three-dimensional (x-y-z axes) nanopositioning in applications such as AFM, where the vertical z-axis movement can be generated by the lateral x-y axes scanning. Such x/y-to-z cross-coupling becomes pronounced when the scanning is at large range and/or at high speed. In AFM applications, the coupling-caused position errors, when large, can generate various adverse effects, including large imaging and topography distortions, and damage of the cantilever probe and/or the sample. This paper utilizes the IIC technique to obtain the control input to precisely track the coupling-caused x/y-to-z displacement (with sign-flipped). Then the obtained input is augmented as a feedforward control to the existing feedback control in tapping-mode imaging, resulting in the cancellation of the coupling effect. The proposed approach is illustrated through two exemplary applications in industry, the pole-tip recession examination, and the nanoasperity measurement on hard-disk drive. Experimental results show that the x/y-to-z coupling effect in large-range (20 and 45 microm) tapping-mode imaging at both low to high scan rates (2, 12.2 to 24.4 Hz) can be effectively removed.

  6. Long-range energy transfer in self-assembled quantum dot-DNA cascades

    Science.gov (United States)

    Goodman, Samuel M.; Siu, Albert; Singh, Vivek; Nagpal, Prashant

    2015-11-01

    The size-dependent energy bandgaps of semiconductor nanocrystals or quantum dots (QDs) can be utilized in converting broadband incident radiation efficiently into electric current by cascade energy transfer (ET) between layers of different sized quantum dots, followed by charge dissociation and transport in the bottom layer. Self-assembling such cascade structures with angstrom-scale spatial precision is important for building realistic devices, and DNA-based QD self-assembly can provide an important alternative. Here we show long-range Dexter energy transfer in QD-DNA self-assembled single constructs and ensemble devices. Using photoluminescence, scanning tunneling spectroscopy, current-sensing AFM measurements in single QD-DNA cascade constructs, and temperature-dependent ensemble devices using TiO2 nanotubes, we show that Dexter energy transfer, likely mediated by the exciton-shelves formed in these QD-DNA self-assembled structures, can be used for efficient transport of energy across QD-DNA thin films.The size-dependent energy bandgaps of semiconductor nanocrystals or quantum dots (QDs) can be utilized in converting broadband incident radiation efficiently into electric current by cascade energy transfer (ET) between layers of different sized quantum dots, followed by charge dissociation and transport in the bottom layer. Self-assembling such cascade structures with angstrom-scale spatial precision is important for building realistic devices, and DNA-based QD self-assembly can provide an important alternative. Here we show long-range Dexter energy transfer in QD-DNA self-assembled single constructs and ensemble devices. Using photoluminescence, scanning tunneling spectroscopy, current-sensing AFM measurements in single QD-DNA cascade constructs, and temperature-dependent ensemble devices using TiO2 nanotubes, we show that Dexter energy transfer, likely mediated by the exciton-shelves formed in these QD-DNA self-assembled structures, can be used for efficient

  7. Mathematical Model and Calibration Experiment of a Large Measurement Range Flexible Joints 6-UPUR Six-Axis Force Sensor

    Directory of Open Access Journals (Sweden)

    Yanzhi Zhao

    2016-08-01

    Full Text Available Nowadays improving the accuracy and enlarging the measuring range of six-axis force sensors for wider applications in aircraft landing, rocket thrust, and spacecraft docking testing experiments has become an urgent objective. However, it is still difficult to achieve high accuracy and large measuring range with traditional parallel six-axis force sensors due to the influence of the gap and friction of the joints. Therefore, to overcome the mentioned limitations, this paper proposed a 6-Universal-Prismatic-Universal-Revolute (UPUR joints parallel mechanism with flexible joints to develop a large measurement range six-axis force sensor. The structural characteristics of the sensor are analyzed in comparison with traditional parallel sensor based on the Stewart platform. The force transfer relation of the sensor is deduced, and the force Jacobian matrix is obtained using screw theory in two cases of the ideal state and the state of flexibility of each flexible joint is considered. The prototype and loading calibration system are designed and developed. The K value method and least squares method are used to process experimental data, and in errors of kind Ι and kind II linearity are obtained. The experimental results show that the calibration error of the K value method is more than 13.4%, and the calibration error of the least squares method is 2.67%. The experimental results prove the feasibility of the sensor and the correctness of the theoretical analysis which are expected to be adopted in practical applications.

  8. Mathematical Model and Calibration Experiment of a Large Measurement Range Flexible Joints 6-UPUR Six-Axis Force Sensor.

    Science.gov (United States)

    Zhao, Yanzhi; Zhang, Caifeng; Zhang, Dan; Shi, Zhongpan; Zhao, Tieshi

    2016-08-11

    Nowadays improving the accuracy and enlarging the measuring range of six-axis force sensors for wider applications in aircraft landing, rocket thrust, and spacecraft docking testing experiments has become an urgent objective. However, it is still difficult to achieve high accuracy and large measuring range with traditional parallel six-axis force sensors due to the influence of the gap and friction of the joints. Therefore, to overcome the mentioned limitations, this paper proposed a 6-Universal-Prismatic-Universal-Revolute (UPUR) joints parallel mechanism with flexible joints to develop a large measurement range six-axis force sensor. The structural characteristics of the sensor are analyzed in comparison with traditional parallel sensor based on the Stewart platform. The force transfer relation of the sensor is deduced, and the force Jacobian matrix is obtained using screw theory in two cases of the ideal state and the state of flexibility of each flexible joint is considered. The prototype and loading calibration system are designed and developed. The K value method and least squares method are used to process experimental data, and in errors of kind Ι and kind II linearity are obtained. The experimental results show that the calibration error of the K value method is more than 13.4%, and the calibration error of the least squares method is 2.67%. The experimental results prove the feasibility of the sensor and the correctness of the theoretical analysis which are expected to be adopted in practical applications.

  9. Crystallinity and compositional changes in carbonated apatites: Evidence from 31P solid-state NMR, Raman, and AFM analysis

    Science.gov (United States)

    McElderry, John-David P.; Zhu, Peizhi; Mroue, Kamal H.; Xu, Jiadi; Pavan, Barbara; Fang, Ming; Zhao, Guisheng; McNerny, Erin; Kohn, David H.; Franceschi, Renny T.; Holl, Mark M. Banaszak; Tecklenburg, Mary M. J.; Ramamoorthy, Ayyalusamy; Morris, Michael D.

    2013-10-01

    Solid-state (magic-angle spinning) NMR spectroscopy is a useful tool for obtaining structural information on bone organic and mineral components and synthetic model minerals at the atomic-level. Raman and 31P NMR spectral parameters were investigated in a series of synthetic B-type carbonated apatites (CAps). Inverse 31P NMR linewidth and inverse Raman PO43-ν1 bandwidth were both correlated with powder XRD c-axis crystallinity over the 0.3-10.3 wt% CO32- range investigated. Comparison with bone powder crystallinities showed agreement with values predicted by NMR and Raman calibration curves. Carbonate content was divided into two domains by the 31P NMR chemical shift frequency and the Raman phosphate ν1 band position. These parameters remain stable except for an abrupt transition at 6.5 wt% carbonate, a composition which corresponds to an average of one carbonate per unit cell. This near-binary distribution of spectroscopic properties was also found in AFM-measured particle sizes and Ca/P molar ratios by elemental analysis. We propose that this transition differentiates between two charge-balancing ion-loss mechanisms as measured by Ca/P ratios. These results define a criterion for spectroscopic characterization of B-type carbonate substitution in apatitic minerals.

  10. Quantification of the number of EP3 receptors on a living CHO cell surface by the AFM

    International Nuclear Information System (INIS)

    Kim, Hyonchol; Arakawa, Hideo; Hatae, Noriyuki; Sugimoto, Yukihiko; Matsumoto, Osamu; Osada, Toshiya; Ichikawa, Atsushi; Ikai, Atsushi

    2006-01-01

    The distribution of EP3 receptors on a living cell surface was quantitatively studied by atomic force microscopy (AFM). Green fluorescent protein (GFP) was introduced to the extracellular region of the EP3 receptor on a CHO cell. A microbead was used as a probe to ensure certain contact area, whose surface was coated with anti-GFP antibody. The interactions between the antibodies and GFP molecules on the cell surface were recorded to observe the distribution of the receptors. The result indicated that EP3 receptors were distributed on the CHO cell surface not uniformly but in small patches coincident with immunohistochemical observation. Repeated measurements on the same area of cell surface gave confirmation that it was unlikely that the receptors were extracted from the cell membrane during the experiments. The measurement of single molecular interaction between GFP and the anti-GFP antibody was succeeded on the cell surface using compression-free force spectroscopy. The value of separation work required to break a single molecular pair was estimated to be about 1.5x10 -18 J. The number of EP3 receptor on the CHO cell surface was estimated using this value to be about 1x10 4 under the assumption that the area of the cell surface was about 5000 μm 2 . These results indicated that the number of receptors on a living cell surface could be quantified through the force measurement by the AFM

  11. WE-G-204-02: Utility of a Channelized Hotelling Model Observer Over a Large Range of Angiographic Exposure Levels

    International Nuclear Information System (INIS)

    Fetterly, K; Favazza, C

    2015-01-01

    Purpose: Mathematical model observers provide a figure of merit that simultaneously considers a test object and the contrast, noise, and spatial resolution properties of an imaging system. The purpose of this work was to investigate the utility of a channelized Hotelling model observer (CHO) to assess system performance over a large range of angiographic exposure conditions. Methods: A 4 mm diameter disk shaped, iodine contrast test object was placed on a 20 cm thick Lucite phantom and 1204 image frames were acquired using fixed x-ray beam quality and for several detector target dose (DTD) values in the range 6 to 240 nGy. The CHO was implemented in the spatial domain utilizing 96 Gabor functions as channels. Detectability index (DI) estimates were calculated using the “resubstitution” and “holdout” methods to train the CHO. Also, DI values calculated using discrete subsets of the data were used to estimate a minimally biased DI as might be expected from an infinitely large dataset. The relationship between DI, independently measured CNR, and changes in results expected assuming a quantum limited detector were assessed over the DTD range. Results: CNR measurements demonstrated that the angiography system is not quantum limited due to relatively increasing contamination from electronic noise that reduces CNR for low DTD. Direct comparison of DI versus CNR indicates that the CHO relatively overestimates DI for low DTD and/or underestimates DI values for high DTD. The relative magnitude of the apparent bias error in the DI values was ∼20% over the 40x DTD range investigated. Conclusion: For the angiography system investigated, the CHO can provide a minimally biased figure of merit if implemented over a restricted exposure range. However, bias leads to overestimates of DI for low exposures. This work emphasizes the need to verify CHO model performance during real-world application

  12. WE-G-204-02: Utility of a Channelized Hotelling Model Observer Over a Large Range of Angiographic Exposure Levels

    Energy Technology Data Exchange (ETDEWEB)

    Fetterly, K; Favazza, C [Mayo Clinic, Rochester, MN (United States)

    2015-06-15

    Purpose: Mathematical model observers provide a figure of merit that simultaneously considers a test object and the contrast, noise, and spatial resolution properties of an imaging system. The purpose of this work was to investigate the utility of a channelized Hotelling model observer (CHO) to assess system performance over a large range of angiographic exposure conditions. Methods: A 4 mm diameter disk shaped, iodine contrast test object was placed on a 20 cm thick Lucite phantom and 1204 image frames were acquired using fixed x-ray beam quality and for several detector target dose (DTD) values in the range 6 to 240 nGy. The CHO was implemented in the spatial domain utilizing 96 Gabor functions as channels. Detectability index (DI) estimates were calculated using the “resubstitution” and “holdout” methods to train the CHO. Also, DI values calculated using discrete subsets of the data were used to estimate a minimally biased DI as might be expected from an infinitely large dataset. The relationship between DI, independently measured CNR, and changes in results expected assuming a quantum limited detector were assessed over the DTD range. Results: CNR measurements demonstrated that the angiography system is not quantum limited due to relatively increasing contamination from electronic noise that reduces CNR for low DTD. Direct comparison of DI versus CNR indicates that the CHO relatively overestimates DI for low DTD and/or underestimates DI values for high DTD. The relative magnitude of the apparent bias error in the DI values was ∼20% over the 40x DTD range investigated. Conclusion: For the angiography system investigated, the CHO can provide a minimally biased figure of merit if implemented over a restricted exposure range. However, bias leads to overestimates of DI for low exposures. This work emphasizes the need to verify CHO model performance during real-world application.

  13. Topography induced by sputtering in a magnetic sector instrument: an AFM and SEM study

    International Nuclear Information System (INIS)

    Iacob, E.; Bersani, M.; Lui, A.; Giubertoni, D.; Barozzi, M.; Anderle, M.

    2004-01-01

    Due to the sensitivity, the good depth resolution and the great interest in ultra shallow profile, secondary ion mass spectrometry (SIMS) is one of the prime techniques used in the semiconductor industry. Low impact energy beams are required to profile shallow distributions. Since Cs + beam sputtering can cause morphological artifacts as well as O 2 + beam does, a detailed study is required to understand development and limiting analytical conditions. In this work we analyzed the effect of low energy Cs + primary beam incident at 68 deg. and 78 deg. on different silicon samples. By using atomic force microscopy (AFM) and scanning electron microscopy (SEM) we underline their reliability and correlate the morphological effects to the SIMS analytical parameters and samples characteristics

  14. Hot-Fire Testing of a 1N AF-M315E Thruster

    Science.gov (United States)

    Burnside, Christopher G.; Pedersen, Kevin; Pierce, Charles W.

    2015-01-01

    This hot-fire test continues NASA investigation of green propellant technologies for future missions. To show the potential for green propellants to replace some hydrazine systems in future spacecraft, NASA Marshall Space Flight Center (MSFC) is continuing to embark on hot-fire test campaigns with various green propellant blends. NASA completed a hot-fire test of a 1N AF-M315E monopropellant thruster at the Marshall Space Flight Center in the small altitude test stand located in building 4205. The thruster is a ground test article used for basic performance determination and catalyst studies. The purpose of the hot-fire testing was for performance determination of a 1N size thruster and form a baseline from which to study catalyst performance and life with follow-on testing to be conducted at a later date. The thruster performed as expected. The result of the hot-fire testing are presented in this paper and presentation.

  15. A novel AFM based method for force measurements between individual hair strands

    International Nuclear Information System (INIS)

    Max, Eva; Haefner, Wolfgang; Wilco Bartels, Frank; Sugiharto, Albert; Wood, Claudia; Fery, Andreas

    2010-01-01

    Interactions between hairs and other natural fibers are of broad interest for both applications and fundamental understanding of biological interfaces. We present a novel method, that allows force measurements between individual hair strands. Hair fragments can be laser-cut without altering their surface chemistry. Subsequently, they are glued onto Atomic force microscopy (AFM) cantilevers. This allows carrying out measurements between the hair fragment and surface immobilized hair in a well-defined crossed-cylinder geometry. Both force-distance and friction measurements are feasible. Measurements in air with controlled humidity and in aqueous environment show clear differences which can be explained by the dominating role of capillary interactions in air. Friction is found to be anisotropic, reflecting the fine structure of hair cuticula. While the investigations are focused on the particular example of human hair, we expect that the approach can be extended to other animal/plant fibers and thus offers perspectives for broad spectrum systems.

  16. Statistical length of DNA based on AFM image measured by a computer

    International Nuclear Information System (INIS)

    Chen Xinqing; Qiu Xijun; Zhang Yi; Hu Jun; Wu Shiying; Huang Yibo; Ai Xiaobai; Li Minqian

    2001-01-01

    Taking advantage of image processing technology, the contour length of DNA molecule was measured automatically by a computer. Based on the AFM image of DNA, the topography of DNA was simulated into a curve. Then the DNA length was measured automatically by inserting mode. It was shown that the experimental length of a naturally deposited DNA (180.4 +- 16.4 nm) was well consistent with the theoretical length (185.0 nm). Comparing to other methods, the present approach had advantages of precision and automatism. The stretched DNA was also measured. It present approach had advantages of precision and automatism. The stretched DNA was also measured. It was shown that the experimental length (343.6 +- 20.7 nm) was much longer than the theoretical length (307.0 nm). This result indicated that the stretching process had a distinct effect on the DNA length. However, the method provided here avoided the DNA-stretching effect

  17. AFM topographies of densely packed nanoparticles: a quick way to determine the lateral size distribution by autocorrelation function analysis

    Czech Academy of Sciences Publication Activity Database

    Fekete, Ladislav; Kůsová, Kateřina; Petrák, Václav; Kratochvílová, Irena

    2012-01-01

    Roč. 14, č. 8 (2012), s. 1-10 ISSN 1388-0764 R&D Projects: GA AV ČR KAN200100801; GA TA ČR TA01011165; GA ČR(CZ) GAP304/10/1951; GA ČR GPP204/12/P235 Institutional research plan: CEZ:AV0Z10100520 Keywords : lateral grain size distribution * AFM * autocorrelation function * nanodiamond Subject RIV: BM - Solid Matter Physics ; Magnetism Impact factor: 2.175, year: 2012

  18. Large scale solvothermal synthesis and a strategy to obtain stable Langmuir–Blodgett film of CoFe2O4 nanoparticles

    International Nuclear Information System (INIS)

    Thampi, Arya; Babu, Keerthi; Verma, Seema

    2013-01-01

    Highlights: • Large scale, monodisperse CoFe 2 O 4 nanoparticles by solvothermal route. • LB technique to obtain stable film of CoFe 2 O 4 nanoparticles over a large area. • Hydrophobicity of substrate was enhanced utilizing LB films of cadmium arachidate. • P–A isotherm and AFM cross sectional height profile analysis confirms stability. • Large scale organization of nanoparticles for surface pressure higher than 15 mN/m. -- Abstract: Nearly monodisperse oleic acid coated cobalt ferrite nanoparticles were synthesized in large scale by a simple solvothermal method utilizing N-methyl 2-Pyrrolidone (NMP) as a high boiling solvent. The magnetic oxide was further investigated by X-ray diffraction (XRD), Fourier transform infrared spectroscopy (FTIR), transmission electron microscopy (TEM), high resolution transmission electron microscopy (HRTEM) and vibrating sample magnetometer (VSM). Langmuir–Blodgett (LB) technique is discussed to obtain a 2D assembly of oleic acid coated CoFe 2 O 4 nanoparticles over a large area. We describe a method to obtain stable, condensed three layers of cadmium arachidate on a piranha treated glass substrate. The hydrophobic surface thus obtained was subsequently used for forming a stable monolayer of oleic acid stabilized cobalt ferrite nanoparticles at the air–water interface. The stability of the LB films at the air–water interface was studied by pressure–area isotherm curves and atomic force microscopy (AFM) cross sectional height profile analysis. 2D organization of the magnetic nanoparticles at different surface pressures was studied by TEM. Preparation of large area LB films of CoFe 2 O 4 nanoparticles is reported for a surface pressure more than 15 mN/m

  19. Amyloid–β peptides time-dependent structural modifications: AFM and voltammetric characterization

    Energy Technology Data Exchange (ETDEWEB)

    Enache, Teodor Adrian; Chiorcea-Paquim, Ana-Maria; Oliveira-Brett, Ana Maria, E-mail: brett@ci.uc.pt

    2016-07-05

    The human amyloid beta (Aβ) peptides, Aβ{sub 1-40} and Aβ{sub 1-42}, structural modifications, from soluble monomers to fully formed fibrils through intermediate structures, were investigated, and the results were compared with those obtained for the inverse Aβ{sub 40-1} and Aβ{sub 42-1}, mutant Aβ{sub 1-40}Phe{sup 10} and Aβ{sub 1-40}Nle{sup 35}, and rat Aβ{sub 1-40}Rat peptide sequences. The aggregation was followed at a slow rate, in chloride free media and room temperature, and revealed to be a sequence-structure process, dependent on the physicochemical properties of each Aβ peptide isoforms, and occurring at different rates and by different pathways. The fibrilization process was investigated by atomic force microscopy (AFM), via changes in the adsorption morphology from: (i) initially random coiled structures of ∼0.6 nm height, corresponding to the Aβ peptide monomers in random coil or in α-helix conformations, to (ii) aggregates and protofibrils of 1.5–6.0 nm height and (iii) two types of fibrils, corresponding to the Aβ peptide in a β-sheet configuration. The reactivity of the carbon electrode surface was considered. The hydrophobic surface induced rapid changes of the Aβ peptide conformations, and differences between the adsorbed fibrils, formed at the carbon surface (beaded, thin, <2.0 nm height) or in solution (long, smooth, thick, >2.0 nm height), were detected. Differential pulse voltammetry showed that, according to their primary structure, the Aβ peptides undergo oxidation in one or two steps, the first step corresponding to the tyrosine amino acids oxidation, and the second one to the histidine and methionine amino acids oxidation. The fibrilization process was electrochemically detected via the decrease of the Aβ peptide oxidation peak currents that occurred in a time dependent manner. - Highlights: • The Aβ peptide fibrilization process was followed by AFM and DP voltammetry. • The human Aβ{sub 1-40} and Aβ{sub 1

  20. Evidence for large compositional ranges in coeval melts erupted from Kīlauea's summit reservoir: Chapter 7

    Science.gov (United States)

    Helz, Rosalind T.; Clague, David A.; Mastin, Larry G.; Rose, Timothy R.; Carey, Rebecca; Cayol, Valérie; Poland, Michael P.; Weis, Dominique

    2015-01-01

    Petrologic observations on Kīlauea's lavas include abundant microprobe analyses of glasses, which show the range of melts available in Kīlauea's summit reservoir over time. During the past two centuries, compositions of melts erupted within the caldera have been limited to MgO = 6.3–7.5 wt%. Extracaldera lavas of the 1959, 1971, and 1974 eruptions contain melts with up to 10.2, 8.9, and 9.2 wt% MgO, respectively, and the 1924 tephra contains juvenile Pele's tears with up to 9.1 wt% MgO. Melt compositions from explosive deposits at Kīlauea, including the Keanakāko‘i (A.D. 1500–1800), Kulanaokuaiki (A.D. 400–1000), and Pāhala (10–25 ka) tephra units, show large ranges of MgO contents. The range of melt MgO is 6.5–11.0 wt% for the Keanakāko‘i; the Kulanaokuaiki extends to 12.5% MgO and the Pāhala Ash includes rare shards with 13–14.5% MgO. The frequency distributions for MgO in the Keanakāko‘i and Kulanaokuaiki glasses are bimodal, suggesting preferential magma storage at two different depths. Kīlauea's summit reservoir contains melts ranging from 6.5 to at least 11.0 wt% MgO, and such melts were available for sampling near instantaneously and repeatedly over centuries. More magnesian melts are inferred to have risen directly from greater depth.

  1. EV range sensitivity analysis

    Energy Technology Data Exchange (ETDEWEB)

    Ostafew, C. [Azure Dynamics Corp., Toronto, ON (Canada)

    2010-07-01

    This presentation included a sensitivity analysis of electric vehicle components on overall efficiency. The presentation provided an overview of drive cycles and discussed the major contributors to range in terms of rolling resistance; aerodynamic drag; motor efficiency; and vehicle mass. Drive cycles that were presented included: New York City Cycle (NYCC); urban dynamometer drive cycle; and US06. A summary of the findings were presented for each of the major contributors. Rolling resistance was found to have a balanced effect on each drive cycle and proportional to range. In terms of aerodynamic drive, there was a large effect on US06 range. A large effect was also found on NYCC range in terms of motor efficiency and vehicle mass. figs.

  2. Influence of vegetation physiognomy, elevation and fire frequency on medium and large mammals in two protected areas of the Espinhaço Range

    Directory of Open Access Journals (Sweden)

    Fernando Ferreira de Pinho

    2017-05-01

    Full Text Available ABSTRACT The objectives of this study were to determine the richness of medium and large mammal species in two protected areas of the Espinhaço Mountain Range, state of Minas Gerais, Brazil; and to investigate the factors affecting the occurrence of those species. To accomplish that we placed 49 camera traps activated by heat and motion at Rio Preto State Park (RPSP and 48 at Sempre Vivas National Park (SVNP. We also collected data on three environmental variables: vegetation physiognomy, elevation and wildfire frequency, to evaluate the influence of these factors on species richness and use intensity (inferred from camera trap detection rate by large mammals. We recorded 23 large mammal species in the two parks combined. The lowest species richness was found at the rupestrian habitat of RPSP, and in the open grasslands of SVNP. The forest and savannah physiognomies were used more intensively by large mammals. Species richness was higher and use was greater at lower elevations of RPSP. In SVNP, fire frequency did not affect species richness or use intensity. The savannah habitat had very similar richness compared to the forests of the two protected areas. The high species richness and use intensity observed in these forest habitats highlights the importance of riparian environments in the Cerrado biome. The highest species richness and use intensity observed at low elevation follows patterns found in the literature, probably due to variation in the vegetation, which results in greater resource availability. Although rupestrian habitats at high elevations of the Espinhaço Range are known to have a high degree of endemism for some taxa, large mammal richness and use were not high in this habitat. These results indicate that the protection of native vegetation at lower elevations is crucial for the long-term conservation of large mammals in the Espinhaço Range.

  3. Adsorption characteristics of Au nanoparticles onto poly(4-vinylpyridine) surface revealed by QCM, AFM, UV/vis, and Raman scattering spectroscopy.

    Science.gov (United States)

    Kim, Kwan; Ryoo, Hyunwoo; Lee, Yoon Mi; Shin, Kuan Soo

    2010-02-15

    In this work, we report that the adsorption and aggregation processes of Au nanoparticles on a polymer surface can be monitored by means of surface-enhanced Raman scattering (SERS) spectroscopy. Specifically, we were able to analyze the adsorption process of citrate-stabilized Au nanoparticles onto a film of poly(4-vinylpyridine) (P4VP) by taking a series of SERS spectra, during the self-assembly of Au nanoparticles onto the polymer film. In order to better analyze the SERS spectra, we separately conducted quartz crystal microbalance (QCM), UV/vis spectroscopy, and atomic force microscope (AFM) measurements. The adsorption kinetics revealed by QCM under the in situ conditions was in fair agreement with that determined by the ex situ AFM measurement. The number of Au nanoparticles adsorbed on P4VP increased almost linearly with time: 265 Au nanoparticles per 1microm(2) were adsorbed on the P4VP film after 6h of immersion. The SERS signal measured in the ex situ condition showed a more rapid increase than that of QCM; however, its increasing pattern was quite similar to that of UV/vis absorbance at longer wavelengths, suggesting that Au nanoparticles actually became agglomerated on P4VP. Copyright 2009 Elsevier Inc. All rights reserved.

  4. Electrochemical and AFM characterization on gold and carbon electrodes of a high redox potential laccase from Fusarium proliferatum.

    Science.gov (United States)

    González Arzola, K; Gimeno, Y; Arévalo, M C; Falcón, M A; Hernández Creus, A

    2010-08-01

    The redox potential of the T1 copper site of laccase from Fusarium proliferatum was determined by titration to be about 510 mV vs. SCE (750 mV vs. NHE), which makes it a high redox potential enzyme. Anaerobic electron transfer reactions between laccase and carbon and gold electrodes were detected, both in solution and when the enzyme was adsorbed on these surfaces. In solution, a single high-potential signal (660 mV vs. SCE) was recorded at the carbon surfaces, attributable to the T1 copper site of the enzyme. However, a well-defined oxidative process at about 660 mV and an anodic wave at 350 mV vs. SCE were recorded at the gold electrode, respectively associated with the T1 and T2 copper sites. Laccase-modified carbon electrodes behaved analogously when the enzyme was in solution, unlike laccase adsorbed on gold, which showed only a low-potential signal. Laccase molecules were successfully imaged by AFM; obtaining a thick compact stable film on Au(111), and large aggregates forming a complex network of small branches leaving voids on the HOPG surface. Laccase-modified carbon electrodes retained significant enzymatic activity, efficiently oxidising violuric acid and reducing molecular oxygen. Explanations are proposed for how protein-film organisation affects the electrode function. Copyright (c) 2009 Elsevier B.V. All rights reserved.

  5. Molecular shape and binding force of Mycoplasma mobile's leg protein Gli349 revealed by an AFM study

    Energy Technology Data Exchange (ETDEWEB)

    Lesoil, Charles [Department of Life Science, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology, Nagatsutacho 4259, Midori-ku, Yokohama 226-8501 (Japan); Nonaka, Takahiro [Department of Biology, Graduate School of Science, Osaka City University, Sumiyoshi-ku, Osaka 558-8585 (Japan); Sekiguchi, Hiroshi; Osada, Toshiya [Department of Life Science, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology, Nagatsutacho 4259, Midori-ku, Yokohama 226-8501 (Japan); Miyata, Makoto [Department of Biology, Graduate School of Science, Osaka City University, Sumiyoshi-ku, Osaka 558-8585 (Japan); Afrin, Rehana [Department of Life Science, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology, Nagatsutacho 4259, Midori-ku, Yokohama 226-8501 (Japan); Biofrontier Center, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology, Nagatsutacho 4259, Midori-ku, Yokohama 226-8501 (Japan); Ikai, Atsushi, E-mail: ikai.a.aa@m.titech.ac.jp [Department of Life Science, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology, Nagatsutacho 4259, Midori-ku, Yokohama 226-8501 (Japan)

    2010-01-15

    Recent studies of the gliding bacteria Mycoplasma mobile have identified a family of proteins called the Gli family which was considered to be involved in this novel and yet fairly unknown motility system. The 349 kDa protein called Gli349 was successfully isolated and purified from the bacteria, and electron microscopy imaging and antibody experiments led to the hypothesis that it acts as the 'leg' of M. mobile, responsible for attachment to the substrate as well as for gliding motility. However, more precise evidence of the molecular shape and function of this protein was required to asses this theory any further. In this study, an atomic force microscope (AFM) was used both as an imaging and a force measurement device to provide new information about Gli349 and its role in gliding motility. AFM images of the protein were obtained revealing a complex structure with both rigid and flexible parts, consistent with previous electron micrographs of the protein. Single-molecular force spectroscopy experiments were also performed, revealing that Gli349 is able to specifically bind to sialyllactose molecules and withstand unbinding forces around 70 pN. These findings strongly support the idea that Gli349 is the 'leg' protein of M. mobile, responsible for binding and also most probably force generation during gliding motility.

  6. AFM, SEM and in situ RHEED study of Cu texture evolution on amorphous carbon by oblique angle vapor deposition

    International Nuclear Information System (INIS)

    Tang, F.; Gaire, C.; Ye, D.-X.; Karabacak, T.; Lu, T.-M.; Wang, G.-C.

    2005-01-01

    The evolution of the crystal orientation of a Cu film grown on an amorphous carbon substrate without intentional heating under 75±6 deg. oblique angle vapor deposition was investigated ex-situ by atomic force microscopy (AFM) and scanning electron microscopy (SEM), and in-situ by reflection high-energy electron diffraction (RHEED). At the initial stage of growth ( ∼100 nm thick) the diffraction pattern started to break symmetrically from the middle of the (111) and (200) rings representing the absence of (111) and (200) planes parallel to the substrate. However, after this transition stage, at the thickness of ∼410 nm, the intensity distribution of diffraction patterns appeared asymmetric about the middle of the rings, which is interpreted as the appearance of a tilted (111) texture. Finally the diffraction patterns developed into separated short arcs and showed only a II-O (two-orientation) texture. By comparing RHEED patterns with the SEM and AFM images of the final film, we argue that the tilted columns having tilted (111) top faces dominate in the later stage of growth. Furthermore, considering the geometry of crystals and shadowing effects, we argue that the vertices of columns having the highest growth velocity normal to the substrate and therefore receiving the maximum flux will dominate the film growth and determine the tilt angle of the texture and the preference of the azimuthal angle orientation

  7. Biophysical analysis of bacterial and viral systems. A shock tube study of bio-aerosols and a correlated AFM/nanosims investigation of vaccinia virus

    Energy Technology Data Exchange (ETDEWEB)

    Gates, Sean Damien [Stanford Univ., CA (United States)

    2013-05-01

    The work presented herein is concerned with the development of biophysical methodology designed to address pertinent questions regarding the behavior and structure of select pathogenic agents. Two distinct studies are documented: a shock tube analysis of endospore-laden bio-aerosols and a correlated AFM/NanoSIMS study of the structure of vaccinia virus.

  8. In Situ AFM Imaging of Microstructural Changes Associated with The Spin Transition in [Fe(Htrz)₂(Trz)](Bf₄) Nanoparticles.

    Science.gov (United States)

    Manrique-Juárez, María D; Suleimanov, Iurii; Hernández, Edna M; Salmon, Lionel; Molnár, Gábor; Bousseksou, Azzedine

    2016-06-30

    Topographic images of [Fe(Htrz)₂(trz)](BF₄) nanoparticles were acquired across the first-order spin transition using variable-temperature atomic force microscopy (AFM) in amplitude modulation mode. These studies revealed a complex morphology of the particles consisting of aggregates of small nanocrystals, which expand, separate and re-aggregate due to the mechanical stress during the spin-state switching events. Both reversible (prompt or slow recovery) and irreversible effects (fatigue) on the particle morphology were evidenced and correlated with the spin crossover properties.

  9. Large scale solvothermal synthesis and a strategy to obtain stable Langmuir–Blodgett film of CoFe{sub 2}O{sub 4} nanoparticles

    Energy Technology Data Exchange (ETDEWEB)

    Thampi, Arya; Babu, Keerthi; Verma, Seema, E-mail: sa.verma@iiserpune.ac.in

    2013-07-05

    Highlights: • Large scale, monodisperse CoFe{sub 2}O{sub 4} nanoparticles by solvothermal route. • LB technique to obtain stable film of CoFe{sub 2}O{sub 4} nanoparticles over a large area. • Hydrophobicity of substrate was enhanced utilizing LB films of cadmium arachidate. • P–A isotherm and AFM cross sectional height profile analysis confirms stability. • Large scale organization of nanoparticles for surface pressure higher than 15 mN/m. -- Abstract: Nearly monodisperse oleic acid coated cobalt ferrite nanoparticles were synthesized in large scale by a simple solvothermal method utilizing N-methyl 2-Pyrrolidone (NMP) as a high boiling solvent. The magnetic oxide was further investigated by X-ray diffraction (XRD), Fourier transform infrared spectroscopy (FTIR), transmission electron microscopy (TEM), high resolution transmission electron microscopy (HRTEM) and vibrating sample magnetometer (VSM). Langmuir–Blodgett (LB) technique is discussed to obtain a 2D assembly of oleic acid coated CoFe{sub 2}O{sub 4} nanoparticles over a large area. We describe a method to obtain stable, condensed three layers of cadmium arachidate on a piranha treated glass substrate. The hydrophobic surface thus obtained was subsequently used for forming a stable monolayer of oleic acid stabilized cobalt ferrite nanoparticles at the air–water interface. The stability of the LB films at the air–water interface was studied by pressure–area isotherm curves and atomic force microscopy (AFM) cross sectional height profile analysis. 2D organization of the magnetic nanoparticles at different surface pressures was studied by TEM. Preparation of large area LB films of CoFe{sub 2}O{sub 4} nanoparticles is reported for a surface pressure more than 15 mN/m.

  10. Analysis the effect of different geometries of AFM's cantilever on the dynamic behavior and the critical forces of three-dimensional manipulation

    International Nuclear Information System (INIS)

    Korayem, Moharam Habibnejad; Saraie, Maniya B.; Saraee, Mahdieh B.

    2017-01-01

    An important challenge when using an atomic force microscope (AFM) is to be able to control the force exerted by the AFM for performing various tasks. Nevertheless, the exerted force is proportional to the deflection of the AFM cantilever, which itself is affected by a cantilever's stiffness coefficient. Many papers have been published so far on the methods of obtaining the stiffness coefficients of AFM cantilevers in 2D; however, a comprehensive model is yet to be presented on 3D cantilever motion. The discrepancies between the equations of the 2D and 3D analysis are due to the number and direction of forces and moments that are applied to a cantilever. Moreover, in the 3D analysis, contrary to the 2D analysis, due to the interaction between the forces and moments applied on a cantilever, its stiffness values cannot be separately expressed for each direction; and instead, a stiffness matrix should be used to correctly derive the relevant equations. In this paper, 3D stiffness coefficient matrices have been obtained for three common cantilever geometries including the rectangular, V-shape and dagger-shape cantilevers. The obtained equations are validated by two methods. In the first approach, the Finite Element Method is combined with the cantilever deflection values computed by using the obtained stiffness matrices. In the second approach, by reducing the problem's parameters, the forces applied on a cantilever along different directions are compared with each other in 2D and 3D cases. Then the 3D manipulation of a stiff nanoparticle is modeled and simulated by using the stiffness matrices obtained for the three cantilever geometries. The obtained results indicate that during the manipulation process, the dagger-shaped and rectangular cantilevers exert the maximum and minimum amounts of forces on the stiff nanoparticle, respectively. Also, by examining the effects of different probe tip geometries, it is realized that a probe tip of cylindrical geometry exerts the

  11. An advanced method to assess the diet of free-ranging large carnivores based on scats.

    Directory of Open Access Journals (Sweden)

    Bettina Wachter

    Full Text Available BACKGROUND: The diet of free-ranging carnivores is an important part of their ecology. It is often determined from prey remains in scats. In many cases, scat analyses are the most efficient method but they require correction for potential biases. When the diet is expressed as proportions of consumed mass of each prey species, the consumed prey mass to excrete one scat needs to be determined and corrected for prey body mass because the proportion of digestible to indigestible matter increases with prey body mass. Prey body mass can be corrected for by conducting feeding experiments using prey of various body masses and fitting a regression between consumed prey mass to excrete one scat and prey body mass (correction factor 1. When the diet is expressed as proportions of consumed individuals of each prey species and includes prey animals not completely consumed, the actual mass of each prey consumed by the carnivore needs to be controlled for (correction factor 2. No previous study controlled for this second bias. METHODOLOGY/PRINCIPAL FINDINGS: Here we use an extended series of feeding experiments on a large carnivore, the cheetah (Acinonyx jubatus, to establish both correction factors. In contrast to previous studies which fitted a linear regression for correction factor 1, we fitted a biologically more meaningful exponential regression model where the consumed prey mass to excrete one scat reaches an asymptote at large prey sizes. Using our protocol, we also derive correction factor 1 and 2 for other carnivore species and apply them to published studies. We show that the new method increases the number and proportion of consumed individuals in the diet for large prey animals compared to the conventional method. CONCLUSION/SIGNIFICANCE: Our results have important implications for the interpretation of scat-based studies in feeding ecology and the resolution of human-wildlife conflicts for the conservation of large carnivores.

  12. Fourier Transform Infrared (FTIR) Spectroscopy, Ultraviolet Resonance Raman (UVRR) Spectroscopy, and Atomic Force Microscopy (AFM) for Study of the Kinetics of Formation and Structural Characterization of Tau Fibrils.

    Science.gov (United States)

    Ramachandran, Gayathri

    2017-01-01

    Kinetic studies of tau fibril formation in vitro most commonly employ spectroscopic probes such as thioflavinT fluorescence and laser light scattering or negative stain transmission electron microscopy. Here, I describe the use of Fourier transform infrared (FTIR) spectroscopy, ultraviolet resonance Raman (UVRR) spectroscopy, and atomic force microscopy (AFM) as complementary probes for studies of tau aggregation. The sensitivity of vibrational spectroscopic techniques (FTIR and UVRR) to secondary structure content allows for measurement of conformational changes that occur when the intrinsically disordered protein tau transforms into cross-β-core containing fibrils. AFM imaging serves as a gentle probe of structures populated over the time course of tau fibrillization. Together, these assays help further elucidate the structural and mechanistic complexity inherent in tau fibril formation.

  13. SNOM and AFM microscopy techniques to study the effect of non-ionizing radiation on the morphological and biochemical properties of human keratinocytes cell line (HaCaT).

    Science.gov (United States)

    Rieti, S; Manni, V; Lisi, A; Giuliani, L; Sacco, D; D'Emilia, E; Cricenti, A; Generosi, R; Luce, M; Grimaldi, S

    2004-01-01

    In this study we have employed atomic force microscopy (AFM) and scanning near-field optical microscopy (SNOM) techniques to study the effect of the interaction between human keratinocytes (HaCaT) and electromagnetic fields at low frequency. HaCaT cells were exposed to a sinusoidal magnetic field at a density of 50 Hz, 1 mT. AFM analysis revealed modification in shape and morphology in exposed cells with an increase in the areas of adhesion between cells. This latter finding was confirmed by SNOM indirect immunofluorescence analysis performed with a fluorescent antibody against the adhesion marker beta4 integrin, which revealed an increase of beta4 integrin segregation in the cell membrane of 50-Hz exposed cells, suggesting that a higher percentage of these cells shows a modified pattern of this adhesion marker.

  14. Detection of aflatoxin M1 in powdered milk and sweetened condensed milk products in several cities in Java with HPLC-fluorescence method

    Science.gov (United States)

    Wijaya, H.; Wardayanie, N. I.; Widjajanti, R.; Silitonga, R. F.

    2018-01-01

    Aflatoxin M1 (AFM1) is a hydroxylated metabolite of aflatoxin B1 (AFB1) produced by lactating animals due to consuming AFB1-contaminated feed. AFM1 can be found in dairy products because it is resistant to heat during processing. This study aimed to detect AFM1 in powdered milk and sweetened condensed milk sold in several cities in Java. The amount of powdered milk sample was 20, while the amount of sweetened condensed milk sample was 16. AFM1 detection in powdered milk and sweetened condensed milk was conducted by HPLC-fluorescence method. The results showed that the concentration of AFM1 in powdered milk ranged from undetectable to 0.549 μg/kg and the highest data (55%) was distributed in concentration range of >0.05 μg/kg - 0.2 μg/kg. On the other hand, AFM1 levels in sweetened condensed milk ranged from undetectable to 0.056 μg/kg and 43.75% data was distributed in concentration range of >0.025 μg/kg - 0.05 μg/kg. All powdered milk and sweetened condensed milk samples have met the maximum level of AFM1 according to Indonesian regulation.

  15. High-resolution and large dynamic range nanomechanical mapping in tapping-mode atomic force microscopy

    International Nuclear Information System (INIS)

    Sahin, Ozgur; Erina, Natalia

    2008-01-01

    High spatial resolution imaging of material properties is an important task for the continued development of nanomaterials and studies of biological systems. Time-varying interaction forces between the vibrating tip and the sample in a tapping-mode atomic force microscope contain detailed information about the elastic, adhesive, and dissipative response of the sample. We report real-time measurement and analysis of the time-varying tip-sample interaction forces with recently introduced torsional harmonic cantilevers. With these measurements, high-resolution maps of elastic modulus, adhesion force, energy dissipation, and topography are generated simultaneously in a single scan. With peak tapping forces as low as 0.6 nN, we demonstrate measurements on blended polymers and self-assembled molecular architectures with feature sizes at 1, 10, and 500 nm. We also observed an elastic modulus measurement range of four orders of magnitude (1 MPa to 10 GPa) for a single cantilever under identical feedback conditions, which can be particularly useful for analyzing heterogeneous samples with largely different material components.

  16. Application of Electron Backscattered Diffraction (EBSD) and Atomic Force Microscopy (AFM) to Determine Texture, Mesotexture, and Grain Boundary Energies in Ceramics

    International Nuclear Information System (INIS)

    Glass, S.J.; Rohrer, G.S.; Saylor, D.M.; Vedula, V.R.

    1999-01-01

    Crystallographic orientations in alumina (Al 2 0 3 ) and magnesium aluminate spinel (MgAl 2 0 4 ) were obtained using electron backscattered diffraction (EBSD) patterns. The texture and mesotexture (grain boundary mis-orientations) were random and no special boundaries were observed. The relative grain boundary energies were determined by thermal groove geometries using atomic force microscopy (AFM) to identify relationships between the grain boundary energies and mis-orientations

  17. Determining the mechanical properties of electrospun poly-ε-caprolactone (PCL) nanofibers using AFM and a novel fiber anchoring technique.

    Science.gov (United States)

    Baker, Stephen R; Banerjee, Soham; Bonin, Keith; Guthold, Martin

    2016-02-01

    Due to its low cost, biocompatibility and slow bioresorption, poly-ε-caprolactone (PCL) continues to be a suitable material for select biomedical engineering applications. We used a combined atomic force microscopy (AFM)/optical microscopy technique to determine key mechanical properties of individual electrospun PCL nanofibers with diameters between 440-1040nm. Compared to protein nanofibers, PCL nanofibers showed much lower adhesion, as they slipped on the substrate when mechanically manipulated. We, therefore, first developed a novel technique to anchor individual PCL nanofibers to micrometer-sized ridges on a substrate, and then mechanically tested anchored nanofibers. When held at constant strain, tensile stress relaxed with fast and slow relaxation times of 1.0±0.3s and 8.8±3.1s, respectively. The total tensile modulus was 62±26MPa, the elastic (non-relaxing) component of the tensile modulus was 53±36MPa. Individual PCL fibers could be stretched elastically (without permanent deformation) to strains of 19-23%. PCL nanofibers are rather extensible; they could be stretched to a strain of at least 98%, and a tensile strength of at least 12MPa, before they slipped off the AFM tip. PCL nanofibers that had aged for over a month at ambient conditions became stiffer and less elastic. Our technique provides accurate nanofiber mechanical data, which are needed to guide construction of scaffolds for cells and other biomedical devices. Copyright © 2015 The Authors. Published by Elsevier B.V. All rights reserved.

  18. UV laser ablation of intraocular lenses: SEM and AFM microscopy examination of the biomaterial surface

    International Nuclear Information System (INIS)

    Spyratou, E.; Asproudis, I.; Tsoutsi, D.; Bacharis, C.; Moutsouris, K.; Makropoulou, M.; Serafetinides, A.A.

    2010-01-01

    Several new materials and patterns are studied for the formation and etching of intraocular lenses (IOLs), in order to improve their optical properties, to reduce the diffractive aberrations and to decrease the incidence of posterior capsular opacification. The aim of this study is to investigate the use of UV (λ = 266 nm) laser pulses to ablate the intraocular lenses materials, and thus to provide an alternative to conventional surface shaping techniques for IOLs fabrication. Ablation experiments were conducted using various polymer substrates of hydrophobic acrylic IOLs and PMMA IOLs. We investigated the ablation efficiency and the morphology of the ablated area by imaging the surface modification with atomic force microscopy (AFM) and scanning electron microscopy (SEM). The morphological appearance of IOL samples reveals the effect of a photochemical and photothermal ablation mechanism.

  19. UV laser ablation of intraocular lenses: SEM and AFM microscopy examination of the biomaterial surface

    Energy Technology Data Exchange (ETDEWEB)

    Spyratou, E., E-mail: ellas5@central.ntua.gr [National Technical University of Athens, School of Applied Mathematical and Physical Sciences, Department of Physics, Zografou Campus, Athens, 15780 (Greece); Asproudis, I. [Department of Ophthalmology, University Hospital of Ioannina, Ioannina, 45110 (Greece); Tsoutsi, D. [Department of Chemistry, University of Ioannina, Ioannina, 45110 (Greece); Bacharis, C.; Moutsouris, K.; Makropoulou, M.; Serafetinides, A.A. [National Technical University of Athens, School of Applied Mathematical and Physical Sciences, Department of Physics, Zografou Campus, Athens, 15780 (Greece)

    2010-02-01

    Several new materials and patterns are studied for the formation and etching of intraocular lenses (IOLs), in order to improve their optical properties, to reduce the diffractive aberrations and to decrease the incidence of posterior capsular opacification. The aim of this study is to investigate the use of UV ({lambda} = 266 nm) laser pulses to ablate the intraocular lenses materials, and thus to provide an alternative to conventional surface shaping techniques for IOLs fabrication. Ablation experiments were conducted using various polymer substrates of hydrophobic acrylic IOLs and PMMA IOLs. We investigated the ablation efficiency and the morphology of the ablated area by imaging the surface modification with atomic force microscopy (AFM) and scanning electron microscopy (SEM). The morphological appearance of IOL samples reveals the effect of a photochemical and photothermal ablation mechanism.

  20. Complex magnetic properties and large magnetocaloric effects in RCoGe (R=Tb, Dy compounds

    Directory of Open Access Journals (Sweden)

    Yan Zhang

    2018-05-01

    Full Text Available Complicated magnetic phase transitions and Large magnetocaloric effects (MCEs in RCoGe (R=Tb, Dy compounds have been reported in this paper. Results show that the TbCoGe compounds have a magnetic phase transition from antiferromagnetic to paramagnetic (AFM-PM at TN∼16 K, which is close to the value reported by neutron diffraction. The DyCoGe compound undergoes complicated phase changes from 2 K up to 300 K. The peak at 10 K displays a phase transition from antiferromagnetic to ferromagnetic (AFM-FM. In particular, a significant ferromagnetic to paramagnetic (FM-PM phase transition was found at the temperature as high as 175 K and the cusp becomes more abrupt with the magnetic field increasing from 0.01 T to 0.1 T. The maximum value of magnetic entropy change of TbCoGe and DyCoGe compounds achieve 14.5 J/kg K and 11.5 J/kg K respectively for a field change of 0-5 T. Additionally, the correspondingly considerable refrigerant capacity value of 260 J/kg and 242 J/kg are also obtained respectively, suggesting that both TbCoGe and DyCoGe compounds could be considered as good candidates for low temperature magnetic refrigerant.

  1. The effect of large decoherence on mixing time in continuous-time quantum walks on long-range interacting cycles

    Energy Technology Data Exchange (ETDEWEB)

    Salimi, S; Radgohar, R, E-mail: shsalimi@uok.ac.i, E-mail: r.radgohar@uok.ac.i [Faculty of Science, Department of Physics, University of Kurdistan, Pasdaran Ave, Sanandaj (Iran, Islamic Republic of)

    2010-01-28

    In this paper, we consider decoherence in continuous-time quantum walks on long-range interacting cycles (LRICs), which are the extensions of the cycle graphs. For this purpose, we use Gurvitz's model and assume that every node is monitored by the corresponding point-contact induced by the decoherence process. Then, we focus on large rates of decoherence and calculate the probability distribution analytically and obtain the lower and upper bounds of the mixing time. Our results prove that the mixing time is proportional to the rate of decoherence and the inverse of the square of the distance parameter (m). This shows that the mixing time decreases with increasing range of interaction. Also, what we obtain for m = 0 is in agreement with Fedichkin, Solenov and Tamon's results [48] for cycle, and we see that the mixing time of CTQWs on cycle improves with adding interacting edges.

  2. Improving the scanning speed of atomic force microscopy at the scanning range of several tens of micrometers

    International Nuclear Information System (INIS)

    Wang, Yanyan; Hu, Xiaodong; Xu, Linyan; Hu, Xiaotang

    2013-01-01

    The atomic force microscope (AFM) is a powerful instrument which can measure the surface of samples at the nanoscale. The resonance of the scanner in xy directions, and the feedback control in the z direction are two major sources of image distortion at high scan speed. In order to improve the scanning speed of the AFM, a low-cost and easy method, which includes sinusoidal scans in the fast scan direction, and an intelligent fuzzy controller in the z direction, is proposed in this paper. The use of a single-frequency driving signal in the fast scan direction allows the scanner to move at a higher speed without exciting its mechanical resonance. The intelligent fuzzy controller automatically selects appropriate PI parameters through the analysis of the tracking errors, thus improving the dynamic tracking performance of the z scanner. The development and functioning of the sinusoidal fast scans and the intelligent fuzzy controller are demonstrated, as well as how this approach significantly achieves faster scans and a higher resolution AFM imaging. -- Highlights: ► The sinusoidal scan and the intelligent controller are used to improve AFM's rate. ► A new method is raised to overcome the nonlinearity caused by the sinusoidal scan. ► A new controller is proposed to improve the performance of the vertical direction.

  3. Large diurnal temperature range increases bird sensitivity to climate change

    NARCIS (Netherlands)

    Briga, Michael; Verhulst, Simon

    2015-01-01

    Climate variability is changing on multiple temporal scales, and little is known of the consequences of increases in short-term variability, particularly in endotherms. Using mortality data with high temporal resolution of zebra finches living in large outdoor aviaries (5 years, 359.220 bird-days),

  4. Characterization and analysis of weld lines on micro-injection moulded parts using atomic force microscopy (AFM)

    DEFF Research Database (Denmark)

    Tosello, Guido; Gava, Alberto; Hansen, Hans Nørgaard

    2009-01-01

    In recent years plastic moulding techniques, such as injection moulding, have been developed to fulfil the needs of micro-components fabrication. Micro-injection moulding (SLIM) is the process which enables the mass production of polymer micro-systems such as micro-mechanical parts, micro...... the two original flows will generate and a weld line is formed on the surface of the micro-moulded part. This phenomenon has to be avoided or at least reduced, since in the weld line area the mechanical properties are poorer than in the bulk part, creating strength problems on the final part. Although...... injection moulding parameters on the weld lines' dimensions is presented, using an atomic force microscope (AFM). Depth and width of weld lines were chosen as parameters to be optimized....

  5. Higher order moments of the matter distribution in scale-free cosmological simulations with large dynamic range

    Science.gov (United States)

    Lucchin, Francesco; Matarrese, Sabino; Melott, Adrian L.; Moscardini, Lauro

    1994-01-01

    We calculate reduced moments (xi bar)(sub q) of the matter density fluctuations, up to order q = 5, from counts in cells produced by particle-mesh numerical simulations with scale-free Gaussian initial conditions. We use power-law spectra P(k) proportional to k(exp n) with indices n = -3, -2, -1, 0, 1. Due to the supposed absence of characteristic times or scales in our models, all quantities are expected to depend on a single scaling variable. For each model, the moments at all times can be expressed in terms of the variance (xi bar)(sub 2), alone. We look for agreement with the hierarchical scaling ansatz, according to which ((xi bar)(sub q)) proportional to ((xi bar)(sub 2))(exp (q - 1)). For n less than or equal to -2 models, we find strong deviations from the hierarchy, which are mostly due to the presence of boundary problems in the simulations. A small, residual signal of deviation from the hierarchical scaling is however also found in n greater than or equal to -1 models. The wide range of spectra considered and the large dynamic range, with careful checks of scaling and shot-noise effects, allows us to reliably detect evolution away from the perturbation theory result.

  6. Specific binding of large aggregates of amphiphilic molecules to the respective antibodies.

    Science.gov (United States)

    Nabok, Alexei; Tsargorodskaya, Anna; Holloway, Alan; Starodub, Nikolay F; Demchenko, Anna

    2007-07-31

    The Binding of nonylphenol to respective antibodies immobilized on solid substrates was studied with the methods of total internal reflection ellipsometry (TIRE) and QCM (quartz crystal microbalance) impedance spectroscopy. The binding reaction was proved to be highly specific having an association constant of KA=1.6x10(6) mol(-1) L and resulted in an increase in both the adsorbed layer thickness of 23 nm and the added mass of 18.3 microg/cm2 at saturation. The obtained responses of both TIRE and QCM methods are substantially higher than anticipated for the immune binding of single molecules of nonylphenol. The mechanism of binding of large aggregates of nonylphenol was suggested instead. Modeling of the micelle of amphiphilic nonylphenol molecules in aqueous solutions yielded a micelle size of about 38 nm. The mechanism of binding of large molecular aggregates to respective antibodies can be extended to other hydrophobic low-molecular-weight toxins such as T-2 mycotoxin. The formation of large molecular aggregates of nonylphenol and T-2 mycotoxin molecules on the surface was proved by the AFM study.

  7. Quantification of the electrostatic forces involved in the directed assembly of colloidal nanoparticles by AFM nanoxerography.

    Science.gov (United States)

    Palleau, E; Sangeetha, N M; Ressier, L

    2011-08-12

    Directed assembly of 10 nm dodecanethiol stabilized silver nanoparticles in hexane and 14 nm citrate stabilized gold nanoparticles in ethanol was performed by AFM nanoxerography onto charge patterns of both polarities written into poly(methylmethacrylate) thin films. The quasi-neutral silver nanoparticles were grafted on both positive and negative charge patterns while the negatively charged gold nanoparticles were selectively deposited on positive charge patterns only. Numerical simulations were conducted to quantify the magnitude, direction and spatial range of the electrophoretic and dielectrophoretic forces exerted by the charge patterns on these two types of nanoparticles in suspension taken as models. The simulations indicate that the directed assembly of silver nanoparticles on both charge patterns is due to the predominant dielectrophoretic forces, while the selective assembly of gold nanoparticles only on positive charge patterns is due to the predominant electrophoretic forces. The study also suggests that the minimum surface potential of charge patterns required for obtaining effective nanoparticle assembly depends strongly on the charge and polarizability of the nanoparticles and also on the nature of the dispersing solvent. Attractive electrostatic forces of about 2 × 10( - 2) pN in magnitude just above the charged surface appear to be sufficient to trap silver nanoparticles in hexane onto charge patterns and the value is about 2 pN for gold nanoparticles in ethanol, under the present experimental conditions. The numerical simulations used in this work to quantify the electrostatic forces operating in the directed assembly of nanoparticles from suspensions onto charge patterns can easily be extended to any kind of colloid and serve as an effective tool for a better comprehension and prediction of liquid-phase nanoxerography processes.

  8. Quantification of the electrostatic forces involved in the directed assembly of colloidal nanoparticles by AFM nanoxerography

    International Nuclear Information System (INIS)

    Palleau, E; Sangeetha, N M; Ressier, L

    2011-01-01

    Directed assembly of 10 nm dodecanethiol stabilized silver nanoparticles in hexane and 14 nm citrate stabilized gold nanoparticles in ethanol was performed by AFM nanoxerography onto charge patterns of both polarities written into poly(methylmethacrylate) thin films. The quasi-neutral silver nanoparticles were grafted on both positive and negative charge patterns while the negatively charged gold nanoparticles were selectively deposited on positive charge patterns only. Numerical simulations were conducted to quantify the magnitude, direction and spatial range of the electrophoretic and dielectrophoretic forces exerted by the charge patterns on these two types of nanoparticles in suspension taken as models. The simulations indicate that the directed assembly of silver nanoparticles on both charge patterns is due to the predominant dielectrophoretic forces, while the selective assembly of gold nanoparticles only on positive charge patterns is due to the predominant electrophoretic forces. The study also suggests that the minimum surface potential of charge patterns required for obtaining effective nanoparticle assembly depends strongly on the charge and polarizability of the nanoparticles and also on the nature of the dispersing solvent. Attractive electrostatic forces of about 2 x 10 -2 pN in magnitude just above the charged surface appear to be sufficient to trap silver nanoparticles in hexane onto charge patterns and the value is about 2 pN for gold nanoparticles in ethanol, under the present experimental conditions. The numerical simulations used in this work to quantify the electrostatic forces operating in the directed assembly of nanoparticles from suspensions onto charge patterns can easily be extended to any kind of colloid and serve as an effective tool for a better comprehension and prediction of liquid-phase nanoxerography processes.

  9. Super-ranging. A new ranging strategy in European badgers.

    Directory of Open Access Journals (Sweden)

    Aoibheann Gaughran

    Full Text Available We monitored the ranging of a wild European badger (Meles meles population over 7 years using GPS tracking collars. Badger range sizes varied seasonally and reached their maximum in June, July and August. We analysed the summer ranging behaviour, using 83 home range estimates from 48 individuals over 6974 collar-nights. We found that while most adult badgers (males and females remained within their own traditional social group boundaries, several male badgers (on average 22% regularly ranged beyond these traditional boundaries. These adult males frequently ranged throughout two (or more social group's traditional territories and had extremely large home ranges. We therefore refer to them as super-rangers. While ranging across traditional boundaries has been recorded over short periods of time for extraterritorial mating and foraging forays, or for pre-dispersal exploration, the animals in this study maintained their super-ranges from 2 to 36 months. This study represents the first time such long-term extra-territorial ranging has been described for European badgers. Holding a super-range may confer an advantage in access to breeding females, but could also affect local interaction networks. In Ireland & the UK, badgers act as a wildlife reservoir for bovine tuberculosis (TB. Super-ranging may facilitate the spread of disease by increasing both direct interactions between conspecifics, particularly across social groups, and indirect interactions with cattle in their shared environment. Understanding super-ranging behaviour may both improve our understanding of tuberculosis epidemiology and inform future control strategies.

  10. Large-scale range collapse of Hawaiian forest birds under climate change and the need 21st century conservation options

    Science.gov (United States)

    Fortini, Lucas B.; Vorsino, Adam E.; Amidon, Fred A.; Paxton, Eben H.; Jacobi, James D.

    2015-01-01

    Hawaiian forest birds serve as an ideal group to explore the extent of climate change impacts on at-risk species. Avian malaria constrains many remaining Hawaiian forest bird species to high elevations where temperatures are too cool for malaria's life cycle and its principal mosquito vector. The impact of climate change on Hawaiian forest birds has been a recent focus of Hawaiian conservation biology, and has centered on the links between climate and avian malaria. To elucidate the differential impacts of projected climate shifts on species with known varying niches, disease resistance and tolerance, we use a comprehensive database of species sightings, regional climate projections and ensemble distribution models to project distribution shifts for all Hawaiian forest bird species. We illustrate that, under a likely scenario of continued disease-driven distribution limitation, all 10 species with highly reliable models (mostly narrow-ranged, single-island endemics) are expected to lose >50% of their range by 2100. Of those, three are expected to lose all range and three others are expected to lose >90% of their range. Projected range loss was smaller for several of the more widespread species; however improved data and models are necessary to refine future projections. Like other at-risk species, Hawaiian forest birds have specific habitat requirements that limit the possibility of range expansion for most species, as projected expansion is frequently in areas where forest habitat is presently not available (such as recent lava flows). Given the large projected range losses for all species, protecting high elevation forest alone is not an adequate long-term strategy for many species under climate change. We describe the types of additional conservation actions practitioners will likely need to consider, while providing results to help with such considerations.

  11. Analysis the effect of different geometries of AFM's cantilever on the dynamic behavior and the critical forces of three-dimensional manipulation

    Energy Technology Data Exchange (ETDEWEB)

    Korayem, Moharam Habibnejad, E-mail: hkorayem@iust.ac.ir; Saraie, Maniya B.; Saraee, Mahdieh B.

    2017-04-15

    An important challenge when using an atomic force microscope (AFM) is to be able to control the force exerted by the AFM for performing various tasks. Nevertheless, the exerted force is proportional to the deflection of the AFM cantilever, which itself is affected by a cantilever's stiffness coefficient. Many papers have been published so far on the methods of obtaining the stiffness coefficients of AFM cantilevers in 2D; however, a comprehensive model is yet to be presented on 3D cantilever motion. The discrepancies between the equations of the 2D and 3D analysis are due to the number and direction of forces and moments that are applied to a cantilever. Moreover, in the 3D analysis, contrary to the 2D analysis, due to the interaction between the forces and moments applied on a cantilever, its stiffness values cannot be separately expressed for each direction; and instead, a stiffness matrix should be used to correctly derive the relevant equations. In this paper, 3D stiffness coefficient matrices have been obtained for three common cantilever geometries including the rectangular, V-shape and dagger-shape cantilevers. The obtained equations are validated by two methods. In the first approach, the Finite Element Method is combined with the cantilever deflection values computed by using the obtained stiffness matrices. In the second approach, by reducing the problem's parameters, the forces applied on a cantilever along different directions are compared with each other in 2D and 3D cases. Then the 3D manipulation of a stiff nanoparticle is modeled and simulated by using the stiffness matrices obtained for the three cantilever geometries. The obtained results indicate that during the manipulation process, the dagger-shaped and rectangular cantilevers exert the maximum and minimum amounts of forces on the stiff nanoparticle, respectively. Also, by examining the effects of different probe tip geometries, it is realized that a probe tip of cylindrical geometry

  12. SFG and AFM Studies of Polymer Surface Monolayers

    Science.gov (United States)

    Somorjai, Gabor A.

    2003-03-01

    Sum frequency generation vibrational spectroscopy and atomic force microscopy techniques were utilized to study the structure and composition of polymer surfaces ranging from polyethylene and polypropylene to copolymers of polyurethane and polystyrene. The surface methyl groups aligned perpendicular to the surface above the glass transition temperature of polypropylene. Large side groups such as the phenyl group on polystyrene is also near the surface normal at the polymer-air interface. At the air interface hydrophobic groups are dominant on the polymer surface while at solid-water interface hydrophilic groups segregate to the surface. Minimizing surface energy is the cause of readjusting the surface composition at polymer-water interfaces as compared to polymer-air interfaces. Upon stretching the soft component of two-component polymer systems segregates to the surface and both the surface structure and the surface composition undergo reversible or irreversible changes depending on the magnitude of the stretch. Since the heart beat forces bio-polymers to stretch over 40 million times a year the molecular behavior due to stretching has important physiological consequences.

  13. Biaxial direct tensile tests in a large range of strain rates. Results on a ferritic nuclear steel

    Energy Technology Data Exchange (ETDEWEB)

    Albertini, C.; Labibes, K.; Montagnani, M.; Pizzinato, E.V.; Solomos, G.; Viaccoz, B. [Commission of the European Communities, Ispra (Italy). Joint Research Centre

    2000-09-01

    Constitutive equations are usually calibrated only trough the experimental results obtained by means of unixial tests because of the lack of adequate biaxial experimental data especially at high strain rate conditions. These data are however important for the validation of analytical models and also for the predictions of mechanical behaviour of real structures subjected to multiaxial loading by numerical simulations. In this paper some developments are shown concerning biaxial cruciform specimens and different experimental machines allowing biaxial tests in a large range of strain rates. This experimental campaign has also allowed study of the influence of changing the strain paths. Diagrams of equivalent stress versus straining direction and also equivalent plastic fracture strain versus straining direction are shown. (orig.)

  14. Journal of Chemical Sciences | Indian Academy of Sciences

    Indian Academy of Sciences (India)

    We describe a novel process for the production of nanoparticles of Cu, Ag, Fe and Al which involves exploding their respective wires, triggered by large current densities in the wires. The particles are characterised by X-ray diffraction (XRD) and atomic force microscopy (AFM). Particle sizes in the range 20-100nm were ...

  15. Optical and AFM study of electrostatically assembled films of CdS and ZnS colloid nanoparticles

    International Nuclear Information System (INIS)

    Suryajaya; Nabok, A.; Davis, F.; Hassan, A.; Higson, S.P.J.; Evans-Freeman, J.

    2008-01-01

    CdS and ZnS semiconducting colloid nanoparticles coated with the organic shell, containing either SO 3 - or NH 2 + groups, were prepared using the aqueous phase synthesis. The multilayer films of CdS (or ZnS) were deposited onto glass, quartz and silicon substrates using the technique of electrostatic self-assembly. The films produced were characterized with UV-vis spectroscopy, spectroscopic ellipsometry and atomic force microscopy. A substantial blue shift of the main absorption band with respect to the bulk materials was found for both CdS and ZnS films. The Efros equation in the effective mass approximation (EMA) theoretical model allowed the evaluation of the nanoparticle radius of 1.8 nm, which corresponds well to the ellipsometry results. AFM shows the formation of larger aggregates of nanoparticles on solid surfaces

  16. Large magnetoelectric coupling in magnetically short-range ordered Bi₅Ti₃FeO₁₅ film.

    Science.gov (United States)

    Zhao, Hongyang; Kimura, Hideo; Cheng, Zhenxiang; Osada, Minoru; Wang, Jianli; Wang, Xiaolin; Dou, Shixue; Liu, Yan; Yu, Jianding; Matsumoto, Takao; Tohei, Tetsuya; Shibata, Naoya; Ikuhara, Yuichi

    2014-06-11

    Multiferroic materials, which offer the possibility of manipulating the magnetic state by an electric field or vice versa, are of great current interest. However, single-phase materials with such cross-coupling properties at room temperature exist rarely in nature; new design of nano-engineered thin films with a strong magneto-electric coupling is a fundamental challenge. Here we demonstrate a robust room-temperature magneto-electric coupling in a bismuth-layer-structured ferroelectric Bi₅Ti₃FeO₁₅ with high ferroelectric Curie temperature of ~1000 K. Bi₅Ti₃FeO₁₅ thin films grown by pulsed laser deposition are single-phase layered perovskit with nearly (00l)-orientation. Room-temperature multiferroic behavior is demonstrated by a large modulation in magneto-polarization and magneto-dielectric responses. Local structural characterizations by transmission electron microscopy and Mössbauer spectroscopy reveal the existence of Fe-rich nanodomains, which cause a short-range magnetic ordering at ~620 K. In Bi₅Ti₃FeO₁₅ with a stable ferroelectric order, the spin canting of magnetic-ion-based nanodomains via the Dzyaloshinskii-Moriya interaction might yield a robust magneto-electric coupling of ~400 mV/Oe·cm even at room temperature.

  17. Does this range suit me? Range satisfaction of battery electric vehicle users.

    Science.gov (United States)

    Franke, Thomas; Günther, Madlen; Trantow, Maria; Krems, Josef F

    2017-11-01

    User satisfaction is a vital design criterion for sustainable systems. The present research aimed to understand factors relating to individually perceived range satisfaction of battery electric vehicle (BEV) users. Data from a large-scale BEV field trial (N = 72) were analyzed. Apart from an initial drop in range satisfaction, increasing practical experience was related to increased range satisfaction. Classical indicators of users' mobility profiles (daily travel distances) were only weakly related to lower range satisfaction (not significant), after controlling for practical experience and preferred coverage of mobility needs. The regularity/predictability of users' mobility patterns, the percentage of journeys not coverable because of range issues, and users' individual comfortable range accounted for variance in range satisfaction. Finally, range satisfaction was related to key indicators of general BEV acceptance (e.g., purchase intentions). These results underline the complex dynamics involved in individual range satisfaction, as well as its central role for BEV acceptance. Copyright © 2017 Elsevier Ltd. All rights reserved.

  18. Biomechanical Properties of Murine Meniscus Surface via AFM-based Nanoindentation

    Science.gov (United States)

    Li, Qing; Doyran, Basak; Gamer, Laura W.; Lu, X. Lucas; Qin, Ling; Ortiz, Christine; Grodzinsky, Alan J.; Rosen, Vicki; Han, Lin

    2015-01-01

    This study aimed to quantify the biomechanical properties of murine meniscus surface. Atomic force microscopy (AFM)-based nanoindentation was performed on the central region, proximal side of menisci from 6- to 24-week old male C57BL/6 mice using microspherical tips (Rtip ≈ 5 μm) in PBS. A unique, linear correlation between indentation depth, D, and response force, F, was found on menisci from all age groups. This non-Hertzian behavior is likely due to the dominance of tensile resistance by the collagen fibril bundles on meniscus surface that are mostly aligned along the circumferential direction observed on 12-week old menisci. The indentation resistance was calculated as both the effective stiffness, Sind = dF/dD, and the effective modulus, Eind, via the isotropic Hertz model. Values of Sind and Eind were found to depend on indentation rate, suggesting the existence of poro-viscoelasticity. These values do not significantly vary with anatomical sites, lateral versus medial compartments, or mouse age. In addition, Eind of meniscus surface (e.g., 6.1 ± 0.8 MPa for 12 weeks of age, mean ± SEM, n = 13) was found to be significantly higher than those of meniscus surfaces in other species, and of murine articular cartilage surface (1.4 ± 0.1 MPa, n = 6). In summary, these results provided the first direct mechanical knowledge of murine knee meniscus tissues. We expect this understanding to serve as a mechanics-based benchmark for further probing the developmental biology and osteoarthritis symptoms of meniscus in various murine models. PMID:25817332

  19. Scaling law to determine peak forces in tapping-mode AFM experiments on finite elastic soft matter systems.

    Science.gov (United States)

    Guzman, Horacio V

    2017-01-01

    Analytical equations to estimate the peak force will facilitate the interpretation and the planning of amplitude-modulation force microscopy (tapping mode) experiments. A closed-form analytical equation to estimate the tip-sample peak forces while imaging soft materials in liquid environment and within an elastic deformation regime has been deduced. We have combined a multivariate regression method with input from the virial-dissipation equations and Tatara's bidimensional deformation contact mechanics model. The equation enables to estimate the peak force based on the tapping mode observables, probe characteristics and the material properties of the sample. The accuracy of the equation has been verified by comparing it to numerical simulations for the archetypical operating conditions to image soft matter with high spatial resolution in tapping-mode AFM.

  20. Automated image analysis of atomic force microscopy images of rotavirus particles

    International Nuclear Information System (INIS)

    Venkataraman, S.; Allison, D.P.; Qi, H.; Morrell-Falvey, J.L.; Kallewaard, N.L.; Crowe, J.E.; Doktycz, M.J.

    2006-01-01

    A variety of biological samples can be imaged by the atomic force microscope (AFM) under environments that range from vacuum to ambient to liquid. Generally imaging is pursued to evaluate structural features of the sample or perhaps identify some structural changes in the sample that are induced by the investigator. In many cases, AFM images of sample features and induced structural changes are interpreted in general qualitative terms such as markedly smaller or larger, rougher, highly irregular, or smooth. Various manual tools can be used to analyze images and extract more quantitative data, but this is usually a cumbersome process. To facilitate quantitative AFM imaging, automated image analysis routines are being developed. Viral particles imaged in water were used as a test case to develop an algorithm that automatically extracts average dimensional information from a large set of individual particles. The extracted information allows statistical analyses of the dimensional characteristics of the particles and facilitates interpretation related to the binding of the particles to the surface. This algorithm is being extended for analysis of other biological samples and physical objects that are imaged by AFM

  1. Automated image analysis of atomic force microscopy images of rotavirus particles

    Energy Technology Data Exchange (ETDEWEB)

    Venkataraman, S. [Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831 (United States); Department of Electrical and Computer Engineering, University of Tennessee, Knoxville, TN 37996 (United States); Allison, D.P. [Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831 (United States); Department of Biochemistry, Cellular, and Molecular Biology, University of Tennessee, Knoxville, TN 37996 (United States); Molecular Imaging Inc. Tempe, AZ, 85282 (United States); Qi, H. [Department of Electrical and Computer Engineering, University of Tennessee, Knoxville, TN 37996 (United States); Morrell-Falvey, J.L. [Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831 (United States); Kallewaard, N.L. [Vanderbilt University Medical Center, Nashville, TN 37232-2905 (United States); Crowe, J.E. [Vanderbilt University Medical Center, Nashville, TN 37232-2905 (United States); Doktycz, M.J. [Life Sciences Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831 (United States)]. E-mail: doktyczmj@ornl.gov

    2006-06-15

    A variety of biological samples can be imaged by the atomic force microscope (AFM) under environments that range from vacuum to ambient to liquid. Generally imaging is pursued to evaluate structural features of the sample or perhaps identify some structural changes in the sample that are induced by the investigator. In many cases, AFM images of sample features and induced structural changes are interpreted in general qualitative terms such as markedly smaller or larger, rougher, highly irregular, or smooth. Various manual tools can be used to analyze images and extract more quantitative data, but this is usually a cumbersome process. To facilitate quantitative AFM imaging, automated image analysis routines are being developed. Viral particles imaged in water were used as a test case to develop an algorithm that automatically extracts average dimensional information from a large set of individual particles. The extracted information allows statistical analyses of the dimensional characteristics of the particles and facilitates interpretation related to the binding of the particles to the surface. This algorithm is being extended for analysis of other biological samples and physical objects that are imaged by AFM.

  2. Synthesis of Large-Scale Single-Crystalline Monolayer WS2 Using a Semi-Sealed Method

    Directory of Open Access Journals (Sweden)

    Feifei Lan

    2018-02-01

    Full Text Available As a two-dimensional semiconductor, WS2 has attracted great attention due to its rich physical properties and potential applications. However, it is still difficult to synthesize monolayer single-crystalline WS2 at larger scale. Here, we report the growth of large-scale triangular single-crystalline WS2 with a semi-sealed installation by chemical vapor deposition (CVD. Through this method, triangular single-crystalline WS2 with an average length of more than 300 µm was obtained. The largest one was about 405 μm in length. WS2 triangles with different sizes and thicknesses were analyzed by optical microscope and atomic force microscope (AFM. Their optical properties were evaluated by Raman and photoluminescence (PL spectra. This report paves the way to fabricating large-scale single-crystalline monolayer WS2, which is useful for the growth of high-quality WS2 and its potential applications in the future.

  3. Reliable Exfoliation of Large-Area High-Quality Flakes of Graphene and Other Two-Dimensional Materials.

    Science.gov (United States)

    Huang, Yuan; Sutter, Eli; Shi, Norman N; Zheng, Jiabao; Yang, Tianzhong; Englund, Dirk; Gao, Hong-Jun; Sutter, Peter

    2015-11-24

    Mechanical exfoliation has been a key enabler of the exploration of the properties of two-dimensional materials, such as graphene, by providing routine access to high-quality material. The original exfoliation method, which remained largely unchanged during the past decade, provides relatively small flakes with moderate yield. Here, we report a modified approach for exfoliating thin monolayer and few-layer flakes from layered crystals. Our method introduces two process steps that enhance and homogenize the adhesion force between the outermost sheet in contact with a substrate: Prior to exfoliation, ambient adsorbates are effectively removed from the substrate by oxygen plasma cleaning, and an additional heat treatment maximizes the uniform contact area at the interface between the source crystal and the substrate. For graphene exfoliation, these simple process steps increased the yield and the area of the transferred flakes by more than 50 times compared to the established exfoliation methods. Raman and AFM characterization shows that the graphene flakes are of similar high quality as those obtained in previous reports. Graphene field-effect devices were fabricated and measured with back-gating and solution top-gating, yielding mobilities of ∼4000 and 12,000 cm(2)/(V s), respectively, and thus demonstrating excellent electrical properties. Experiments with other layered crystals, e.g., a bismuth strontium calcium copper oxide (BSCCO) superconductor, show enhancements in exfoliation yield and flake area similar to those for graphene, suggesting that our modified exfoliation method provides an effective way for producing large area, high-quality flakes of a wide range of 2D materials.

  4. Assembly of α-synuclein fibrils in nanoscale studied by peptide truncation and AFM

    International Nuclear Information System (INIS)

    Zhang Feng; Lin Xiaojing; Ji Lina; Du Haining; Tang Lin; He Jianhua; Hu Jun; Hu Hongyu

    2008-01-01

    α-Synuclein (α-Syn) fibrils are the major component of Lewy bodies that are closely associated with the pathogenesis of Parkinson's disease, but the mechanism for the fibril assembly remains poorly understood. Here we report using a combination of peptide truncation and atomic force microscopy (AFM) to elucidate the self-assembly and morphology of the α-Syn fibrils. The results show that protease K significantly slims the fibrils from the mean height of ∼6.6 to ∼4.7 nm, whereas chaotropic denaturant urea completely breaks down the fibrils into small particles. The in situ enzymatic digestion also results in thinning of the fibrils, giving rise to some nicks on the fibrils. Moreover, N- or C-terminally truncated α-Syn fragments assemble into thinner filaments with the heights depending on the peptide lengths. A nine-residue peptide corresponding to the homologous GAV-motif sequence can form very thin (∼2.2 nm) but long (>1 μm) filaments. Thus, the central sequence of α-Syn forms a fibrillar core by cross-β-structure that is flanked by two flexible termini, and the orientation of the fibril growth is perpendicular to the β-sheet structures

  5. A Study on the Occurrence of Aflatoxin M1 in Raw and Pasteurized Milk Produced in Rafsanjan, Iran

    Directory of Open Access Journals (Sweden)

    FaAkrami Mohajeri Akrami Mohajeri

    2015-12-01

    Full Text Available Introduction: Aflatoxins, known as causative factors of hepatic and extra-hepatic carcinogenesis within humans, are extremely teratogenic, mutagenic, toxic, and carcinogenic compounds. Materials & Methods: This study was undertaken to determine the occurrence of aflatoxin M1 (AFM1 in 40 raw milk and 47 pasteurized milk samples collected during spring and winter. In order to analyze the samples, the Enzyme-linked Immunosorbent Assay (ELISA procedure was used. The statistical methods used in this study were based on normal confidence intervals and analysis of variance (ANOVA. Results: Aflatoxin M1 was detected in 97.5% of the raw milk ranging from 6.52 to 68.17 ng/l and 95.7% of the pasteurized milk, ranging from 0.8 to 58.13 ng/l. Toxin levels in 10% of the raw milk and 2.1% of the pasteurized milk samples exceeded the Iranian national standard limit i.e. 50 ng/l.  Due to seasonal variations, mean concentration of AFM1 in the samples collected in winter was significantly (P < 0.03 higher than those collected in the summer. Conclusion: Large amount of AFM1 in milk samples might be a potential hazard for the public health. Reducing the levels of AFB1 in animal feedstuffs can be regarded as the initial step to control the transfer of AFM1 to humans.

  6. Validation of SplitVectors Encoding for Quantitative Visualization of Large-Magnitude-Range Vector Fields.

    Science.gov (United States)

    Henan Zhao; Bryant, Garnett W; Griffin, Wesley; Terrill, Judith E; Jian Chen

    2017-06-01

    We designed and evaluated SplitVectors, a new vector field display approach to help scientists perform new discrimination tasks on large-magnitude-range scientific data shown in three-dimensional (3D) visualization environments. SplitVectors uses scientific notation to display vector magnitude, thus improving legibility. We present an empirical study comparing the SplitVectors approach with three other approaches - direct linear representation, logarithmic, and text display commonly used in scientific visualizations. Twenty participants performed three domain analysis tasks: reading numerical values (a discrimination task), finding the ratio between values (a discrimination task), and finding the larger of two vectors (a pattern detection task). Participants used both mono and stereo conditions. Our results suggest the following: (1) SplitVectors improve accuracy by about 10 times compared to linear mapping and by four times to logarithmic in discrimination tasks; (2) SplitVectors have no significant differences from the textual display approach, but reduce cluttering in the scene; (3) SplitVectors and textual display are less sensitive to data scale than linear and logarithmic approaches; (4) using logarithmic can be problematic as participants' confidence was as high as directly reading from the textual display, but their accuracy was poor; and (5) Stereoscopy improved performance, especially in more challenging discrimination tasks.

  7. Characterization of electronic charged states of P-doped Si quantum dots using AFM/Kelvin probe

    International Nuclear Information System (INIS)

    Makihara, Katsunori; Xu, Jun; Ikeda, Mitsuhisa; Murakami, Hideki; Higashi, Seiichiro; Miyazaki, Seiichi

    2006-01-01

    Phosphorous doping to Si quantum dots was performed by a pulse injection of 1% PH 3 diluted with He during the dot formation on thermally grown SiO 2 from thermal decomposition of pure SiH 4 , and electron charging to and discharging from P-doped Si dots were studied to characterize their electronic charged states using a Kelvin probe technique in atomic force microscopy (AFM). The potential change corresponding to the extraction of one electron from each of the P-doped Si dots was observed after applying a tip bias as low as + 0.2 V while for undoped Si dots, with almost the same size as P-doped Si dots, almost the same amount of the potential change was detectable only when the tip bias was increased to ∼ 1 V. It is likely that, for P-doped Si dots, the electron extraction from the conduction band occurs and results in a positively charged state with ionized P donor

  8. AFM-investigation of differently treated Ti-surfaces with respect to their usability for dental implants

    Energy Technology Data Exchange (ETDEWEB)

    Wille, Sebastian; Adelung, Rainer [Funktionale Nanomaterialien, Institut fuer Materialwissenschaft, CAU Kiel Kaiserstr. 2 24143 Kiel (Germany); Yang, Bin [Klinik fuer Zahnaerztliche Prothetik, Propaedeutik und Werkstoffkunde, Arnold-Heller-Strasse 16, 24105 Kiel (Germany); Groessner-Schreiber, Birte [Klinik fuer Zahnerhaltungskunde und Parodontologie, Arnold-Heller-Str. 16, 24105 Kiel (Germany)

    2008-07-01

    Microbial adherence to dental implant surfaces is one initiating step in the formation of plaque and is considered to be an important event in the pathogenesis of peri-implant disease. Besides good connective tissue adhesion in the transmucosal part of an implant, titanium implants exposed to the oral cavity require surface modification to inhibit the adherence of oral bacteria. Surface roughness and chemical composition of the implant surface were found to have a significant impact on plaque formation. The aim of the present study was to examine bacterial adherence of differently modified potential implant surfaces. Therefore the surface roughness was decreased and for example a thin ceramic or composite layer of antibacterial material was deposited on abutment surface by sputtering. We analyze the new surface with AFM to control the roughness. For further characterization contact angle measurements were carried out. Biocompatibility and antibacterial effects will be determined in cooperation with the dental clinic at the University Kiel.

  9. In situ oxidation of zirconium binary alloys by environmental SEM and analysis by AFM, FIB, and TEM

    International Nuclear Information System (INIS)

    Proff, C.; Abolhassani, S.; Dadras, M.M.; Lemaignan, C.

    2010-01-01

    Binary Zr-alloys containing 1%Fe and 1% Ni (large precipitates) and 1% Cr and 0.6% Nb (small precipitates), as well as a pure Zr sample were exposed in situ at 130 Pa water vapour pressure at 415 o C in an environmental SEM. The surface topography and composition of each sample was characterised before in situ experiments, during and after oxidation. After oxidation the surface was characterised by SEM and EDS, AFM and TEM combined with EDS. Focused ion beam was used to prepare cross sections of the metal-oxide interface and for the preparation of TEM thin foils. The oxidation behaviour of precipitates for these alloying elements can be characterised into two large families, those which show a rapid oxidation and those which induce a delayed oxidation in comparison with the Zr-matrix. At 415 o C after 1 h of oxidation for Zr1%Fe and Zr1%Ni, the formation of protrusions could be detected at the surface, being related to underlying SPP in the oxide. On Zr1%Cr and Zr0.6%Nb unoxidised SPPs were observed in the oxide, close to the metal-oxide interface. These SPPs were, however, oxidised close to the outer surface of the oxide. The surface roughness was increased for all materials after in situ oxidation, however, only for Zr1%Fe and Zr1%Ni protrusions appeared on the surface during oxidation. It was subsequently demonstrated that these latter correspond to the position of SPPs. For Zr1%Fe the surface roughness increased more than in the other materials and on these protrusions small iron oxide crystals have been observed at the surface. These observations confirm that Fe has a different behaviour compared to the other SPP forming elements, and it diffuses out to the free surface of the material. These alloying elements being the constituents of the commercial alloys (Fe and Cr for Zircaloy-4; Fe, Cr and Ni for Zircaloy-2 and Nb for all Nb-containing alloys), this study allows to separate their individual influence and can allow a subsequent comparison to the behaviour

  10. Nanoimprint-defined, large-area meta-surfaces for unidirectional optical transmission with superior extinction in the visible-to-infrared range.

    Science.gov (United States)

    Yao, Yuhan; Liu, He; Wang, Yifei; Li, Yuanrui; Song, Boxiang; Wang, Richard P; Povinelli, Michelle L; Wu, Wei

    2016-07-11

    Optical devices with asymmetric transmission have important applications in optical systems, but optical isolators with the modal asymmetry can only be built using magneto-optical or nonlinear materials, as dictated by the Lorentz reciprocity theorem. However, optical devices with the power asymmetry can be achieved by linear materials such as metals and dielectrics. In this paper, we report a large-area, nanoimprint-defined meta-surface (stacked subwavelength gratings) with high-contrast asymmetric transmittance in the visible-to-infrared wavelength range for TM-polarized light. The physical origin of asymmetric transmission through the meta-surface is studied by analyzing the scattering matrix.

  11. „New approaches to atomic force microscope lithography on silicon"

    DEFF Research Database (Denmark)

    Birkelund, Karen; Thomsen, Erik Vilain; Rasmussen, Jan Pihl

    1997-01-01

    We have investigated new approaches to the formation of conducting nanowires on crystalline silicon surfaces using atomic force microscope (AFM) lithography. To increase processing speed and reduce wear of the AFM tip, large-scale structures are formed with a direct laser write setup, while the AFM...

  12. AFM study of microbial colonization and its deleterious effect on 304 stainless steel by Pseudomonas NCIMB 2021 and Desulfovibrio desulfuricans in simulated seawater

    International Nuclear Information System (INIS)

    Yuan, S.J.; Pehkonen, S.O.

    2009-01-01

    The biofilm colonization dynamics of Pseudomonas NCIMB 2021 and Desulfovibrio desulfuricans (ATCC 27774) on 304 stainless steels (304 SS) was evaluated using atomic force microscopy (AFM) in simulated seawater-based media under aerobic and anaerobic conditions. Results showed that the biofilm formed on the coupon surface by the two strains of bacteria increased in the coverage, heterogeneity and thickness with exposure time, thus resulting in the deterioration of the steel substratum underneath the biofilm in the form of pitting corrosion. The depth of pits induced by D. desulfuricans was larger than that by Pseudomonas NCIMB 2021, which was mainly attributed to the enhanced corrosion of 304 SS coupons by the biogenic sulfide ions, as revealed by the results of X-ray photoelectron spectroscopy (XPS) and Tafel polarization curves. AFM was also used to determine cell attachment/detachment processes of the Pseudomonas and D. desulfuricans bacteria on the coupon surface by quantifying the tip-cell interaction forces. The interactive forces between the tip and the bacterial cell surface were considerably smaller than those between the tip and the cell-cell interface due to the accumulation of extra-cellular polymeric substances (EPS) for both strains. Furthermore, the adhesion forces over the Pseudomonas cells were verified to be more attractive than those of D. desulfuricans due to the former being a slime-producer.

  13. Magnetic short-range order in Gd

    International Nuclear Information System (INIS)

    Child, H.R.

    1978-01-01

    The magnetic short-range order in a ferromagnetic, isotopically enriched 160 Gd metal single crystal has been investigated by quasielastic scattering of 81-meV neutrons. Since Gd behaves as an S-state ion in the metal, little anisotropy is expected in its magnetic behavior. However, the data show that there is anisotropic short-range order present over a large temperature interval both above and below T/sub C/. The data have been analyzed in terms of an Ornstein-Zernike Lorentzian form with anisotropic correlation ranges. These correlation ranges as deduced from the observed data behave normally above T/sub C/ but seem to remain constant over a fairly large interval below T/sub C/ before becoming unobservable at lower temperatures. These observations suggest that the magnetic ordering in Gd may be a more complicated phenomenon than first believed

  14. Nonlinear dynamic response of cantilever beam tip during atomic force microscopy (AFM) nanolithography of copper surface

    International Nuclear Information System (INIS)

    Yeh, Y-L; Jang, M-J; Wang, C-C; Lin, Y-P; Chen, K-S

    2008-01-01

    This paper investigates the nonlinear dynamic response of an atomic force microscope (AFM) cantilever beam tip during the nanolithography of a copper (Cu) surface using a high-depth feed. The dynamic motion of the tip is modeled using a combined approach based on Newton's law and empirical observations. The cutting force is determined from experimental observations of the piling height on the Cu surface and the rotation angle of the cantilever beam tip. It is found that the piling height increases linearly with the cantilever beam carrier velocity. Furthermore, the cantilever beam tip is found to execute a saw tooth motion. Both this motion and the shear cutting force are nonlinear. The elastic modulus in the y direction is variable. Finally, the velocity of the cantilever beam tip as it traverses the specimen surface has a discrete characteristic rather than a smooth, continuous profile

  15. Determining the mechanical properties of electrospun poly-ε-caprolactone (PCL) nanofibers using AFM and a novel fiber anchoring technique

    Energy Technology Data Exchange (ETDEWEB)

    Baker, Stephen R.; Banerjee, Soham; Bonin, Keith; Guthold, Martin, E-mail: gutholdm@wfu.edu

    2016-02-01

    Due to its low cost, biocompatibility and slow bioresorption, poly-ε-caprolactone (PCL) continues to be a suitable material for select biomedical engineering applications. We used a combined atomic force microscopy (AFM)/optical microscopy technique to determine key mechanical properties of individual electrospun PCL nanofibers with diameters between 440–1040 nm. Compared to protein nanofibers, PCL nanofibers showed much lower adhesion, as they slipped on the substrate when mechanically manipulated. We, therefore, first developed a novel technique to anchor individual PCL nanofibers to micrometer-sized ridges on a substrate, and then mechanically tested anchored nanofibers. When held at constant strain, tensile stress relaxed with fast and slow relaxation times of 1.0 ± 0.3 s and 8.8 ± 3.1 s, respectively. The total tensile modulus was 62 ± 26 MPa, the elastic (non-relaxing) component of the tensile modulus was 53 ± 36 MPa. Individual PCL fibers could be stretched elastically (without permanent deformation) to strains of 19–23%. PCL nanofibers are rather extensible; they could be stretched to a strain of at least 98%, and a tensile strength of at least 12 MPa, before they slipped off the AFM tip. PCL nanofibers that had aged for over a month at ambient conditions became stiffer and less elastic. Our technique provides accurate nanofiber mechanical data, which are needed to guide construction of scaffolds for cells and other biomedical devices. - Highlights: • Developed technique to anchor single nanofibers on microridges • Determined mechanical properties of e-spun PCL fibers • E-spun PCL fibers have similar mechanical properties as wet fibrin fibers. • E-spun PCL fiber total elastic modulus is ~ 60 MPa. • E-spun PCL fiber extensibility is > 100%.

  16. Gallotannin-Capped Gold Nanoparticles: Green Synthesis and Enhanced Morphology of AFM Images.

    Science.gov (United States)

    Kim, Jaehyung; Yhim, Won Been; Park, Jong-Won; Lee, Sang-Hyeon; Kim, Tae Yoon; Cha, Song-Hyun; Kim, Hyun-Seok; Jang, Hong-Lae; Cho, Miyeon; Park, Youmie; Cho, Seonho

    2016-06-01

    Gold nanoparticles (AuNPs) were synthesized by a green method using a plant secondary metabolite, gallotannin. Gallotannin was used as a reducing and capping agent to convert gold ions into AuNPs for the generation of gallotannin-capped AuNPs (GT-AuNPs). This synthetic route is ecofriendly and eliminates the use of toxic chemical reducing agents. The characteristic surface plasmon resonance of the GT-AuNPs was observed at 536 nm in the UV-visible spectra. The face-centered cubic structure of GT-AuNPs was verified by X-ray diffraction analysis. The majority of the GT-AuNPs had a spherical shape with an average diameter of 15.93 ± 8.60 nm. Fourier transform infrared spectra suggested that the hydroxyl functional groups of gallotannin were involved in the synthesis of GT-AuNPs. The size and shape of nanoparticles can have a crucial impact on their biological, mechanical, and structural properties. Herein, we developed a modified anisotropic diffusion equation to selectively remove nanoscale experimental noise while preserving nanoscale intrinsic geometry information. To demonstrate the performance of the developed method, the ridge and valley lines were plotted by utilizing the principle curvatures. Compared to the original anisotropic diffusion and raw atomic force microscopy (AFM) experimental data, the developed modified anisotropic diffusion shows excellent performance in nanoscale noise removal while preserving the intrinsic aeometry of the nanoparticles.

  17. Novel interplay between high-Tc superconductivity and antiferromagnetism in Tl-Based Six-CuO2-layered cuprates. 205Tl- and 63Cu-NMR probes

    International Nuclear Information System (INIS)

    Mukuda, Hidekazu; Shiki, Nozomu; Kimoto, Naoki; Yashima, Mitsuharu; Kitaoka, Yoshio; Tokiwa, Kazuyasu; Iyo, Akira

    2016-01-01

    We report 63 Cu- and 205 Tl-NMR studies on six-layered (n = 6) high-T c superconducting (SC) cuprate TlBa 2 Ca 5 Cu 6 O 14+δ (Tl1256) with T c ∼ 100 K, which reveal that antiferromagnetic (AFM) order takes place below T N ∼ 170 K. In this compound, four underdoped inner CuO 2 planes [n(IP) = 4] sandwiched by two outer planes (OPs) are responsible for the onset of AFM order, whereas the nearly optimally-doped OPs responsible for the onset of bulk SC. It is pointed out that an increase in the out-of-plane magnetic interaction within an intra-unit-cell causes T N ∼ 45 K for Tl1245 with n(IP) = 3 to increase to ∼170 K for Tl1256 with n(IP) = 4. It is remarkable that the marked increase in T N and the AFM moments for the IPs does not bring about any reduction in T c , since T c ∼ 100 K is maintained for both compounds with nearly optimally doped OP. We highlight the fact that the SC order for n ≥ 5 is mostly dominated by the long-range in-plane SC correlation even in the multilayered structure, which is insensitive to the magnitude of T N and the AFM moments at the IPs or the AFM interaction among the IPs. These results demonstrate a novel interplay between the SC and AFM orders when the charge imbalance between the IPs and OP is significantly large. (author)

  18. AFM review study on pox viruses and living cells.

    Science.gov (United States)

    Ohnesorge, F M; Hörber, J K; Häberle, W; Czerny, C P; Smith, D P; Binnig, G

    1997-10-01

    Single living cells were studied in growth medium by atomic force microscopy at a high--down to one image frame per second--imaging rate over time periods of many hours, stably producing hundreds of consecutive scans with a lateral resolution of approximately 30-40 nm. The cell was held by a micropipette mounted onto the scanner-piezo as shown in Häberle, W., J. K. H. Hörber, and G. Binnig. 1991. Force microscopy on living cells. J. Vac. Sci. Technol. B9:1210-0000. To initiate specific processes on the cell surface the cells had been infected with pox viruses as reported earlier and, most likely, the liberation of a progeny virion by the still-living cell was observed, hence confirming and supporting earlier results (Häberle, W., J. K. H. Hörber, F. Ohnesorge, D. P. E. Smith, and G. Binnig. 1992. In situ investigations of single living cells infected by viruses. Ultramicroscopy. 42-44:1161-0000; Hörber, J. K. H., W. Häberle, F. Ohnesorge, G. Binnig, H. G. Liebich, C. P. Czerny, H. Mahnel, and A. Mayr. 1992. Investigation of living cells in the nanometer regime with the atomic force microscope. Scanning Microscopy. 6:919-930). Furthermore, the pox viruses used were characterized separately by AFM in an aqueous environment down to the molecular level. Quasi-ordered structural details were resolved on a scale of a few nm where, however, image distortions and artifacts due to multiple tip effects are probably involved--just as in very high resolution (small dark spots in the light microscope, that we believed to be the regions in the cell plasma where viruses are assembled; this is known from the literature on electron microscopy on pox-infected cells and referred to there as "virus factories" (e.g., Moss, B. 1986. Replication of pox viruses. In Fundamental Virology, B. N. Fields and D. M. Knape, editors. Raven Press, New York. 637-655). Therefore, we assume that the cells stay alive during imaging, in our experience for approximately 30-45 h p.i.).

  19. Fabrication and atomic force microscopy/friction force microscopy (AFM/FFM) studies of polyacrylamide-carbon nanotubes (PAM-CNTs) copolymer thin films

    International Nuclear Information System (INIS)

    Li Xuefeng; Guan Wenchao; Yan Haibiao; Huang Lan

    2004-01-01

    A novel polyacrylamide-carbon nanotubes (PAM-CNTs) copolymer has been prepared by ultraviolet radiation initiated polymerization. The PAM-CNTs copolymer was characterized by the instruments of Fourier transform infrared spectroscopy, UV-vis absorbance spectra, fluorescence spectra and transmission electron microscope. The morphology and microtribological properties of PAM-CNTs thin films on mica were investigated by atomic force microscopy/friction force microscopy (AFM/FFM). The friction of the films was stable with the change of applied load and the friction coefficient decreased significantly as the CNTs addition. The results show that the rigid rod-like CNTs in polymer would enhance load-bearing and anti-wear properties of the thin films

  20. Atomic force microscope featuring an integrated optical microscope

    NARCIS (Netherlands)

    Putman, C.A.J.; Putman, Constant A.J.; de Grooth, B.G.; van Hulst, N.F.; Greve, Jan

    1992-01-01

    The atomic force microscope (AFM) is used to image the surface of both conductors and nonconductors. Biological specimens constitute a large group of nonconductors. A disadvantage of most AFM's is the fact that relatively large areas of the sample surface have to be scanned to pinpoint a biological