WorldWideScience

Sample records for handheld x-ray fluorescence

  1. Algorithms for a hand-held miniature x-ray fluorescence analytical instrument

    International Nuclear Information System (INIS)

    Elam, W.T.; Newman, D.; Ziemba, F.

    1998-01-01

    The purpose of this joint program was to provide technical assistance with the development of a Miniature X-ray Fluorescence (XRF) Analytical Instrument. This new XRF instrument is designed to overcome the weaknesses of spectrometers commercially available at the present time. Currently available XRF spectrometers (for a complete list see reference 1) convert spectral information to sample composition using the influence coefficients technique or the fundamental parameters method. They require either a standard sample with composition relatively close to the unknown or a detailed knowledge of the sample matrix. They also require a highly-trained operator and the results often depend on the capabilities of the operator. In addition, almost all existing field-portable, hand-held instruments use radioactive sources for excitation. Regulatory limits on such sources restrict them such that they can only provide relatively weak excitation. This limits all current hand-held XRF instruments to poor detection limits and/or long data collection times, in addition to the licensing requirements and disposal problems for radioactive sources. The new XRF instrument was developed jointly by Quantrad Sensor, Inc., the Naval Research Laboratory (NRL), and the Department of Energy (DOE). This report describes the analysis algorithms developed by NRL for the new instrument and the software which embodies them

  2. Hazardous metals in vintage plastic toys measured by a handheld X-ray fluorescence spectrometer.

    Science.gov (United States)

    Miller, Gillian Zaharias; Harris, Zoe E

    2015-01-01

    Over 100 plastic toys from the 1970s and 1980s, both polyvinyl chloride ("vinyl") and nonvinyl, were analyzed in the study described here using a handheld X-ray fluorescence spectrometer to quantify hazardous metal content. A sampling of recent vinyl toys was also tested. The majority of nonvinyl samples were Fisher Price brand toys. The vinyl toys consisted largely of Barbie dolls and other dolls. Overall, lead or cadmium was found in 67% of vintage plastic toys, frequently at concentrations exceeding current U.S. and European limits. Arsenic was detected at levels of concern in 16% of the samples. In the nonvinyl toys, heavy metal content was found to correlate with certain colors of plastic. The likely sources of the detected metals are discussed. None of the contemporary vinyl toys contained detectable cadmium, lead, or arsenic. Given that vintage toys remain in widespread use by children in homes and other locations, the results illuminate a potential source of heavy metal exposure for children.

  3. Provenance study of Gothic paintings from North-East Slovakia by handheld x-ray fluorescence, microscopy and x-ray microdiffraction

    Czech Academy of Sciences Publication Activity Database

    Hradil, David; Hradilová, J.; Bezdička, Petr; Švarcová, Silvie

    2008-01-01

    Roč. 37, č. 4 (2008), s. 376-382 ISSN 0049-8246 R&D Projects: GA ČR(CZ) GA203/07/1324 Institutional research plan: CEZ:AV0Z40320502 Keywords : Gothic paintings * X-ray fluorescence * X-ray microdiffraction Subject RIV: CA - Inorganic Chemistry Impact factor: 1.390, year: 2008

  4. Possible Radiation-Induced Damage to the Molecular Structure of Wooden Artifacts Due to Micro-Computed Tomography, Handheld X-Ray Fluorescence, and X-Ray Photoelectron Spectroscopic Techniques

    Directory of Open Access Journals (Sweden)

    Madalena Kozachuk

    2016-05-01

    Full Text Available This study was undertaken to ascertain whether radiation produced by X-ray photoelectron spectroscopy (XPS, micro-computed tomography (μCT and/or portable handheld X-ray fluorescence (XRF equipment might damage wood artifacts during analysis. Changes at the molecular level were monitored by Fourier transform infrared (FTIR analysis. No significant changes in FTIR spectra were observed as a result of μCT or handheld XRF analysis. No substantial changes in the collected FTIR spectra were observed when XPS analytical times on the order of minutes were used. However, XPS analysis collected over tens of hours did produce significant changes in the FTIR spectra.

  5. Use of handheld X-ray fluorescence spectrometry units for identification of arsenic in treated wood

    Energy Technology Data Exchange (ETDEWEB)

    Block, Colleen N. [University of Miami, Department of Civil, Architectural, and Environmental Engineering, P.O. Box 248294, McArthur Building, Coral Gables, FL 33124-0630 (United States); Shibata, Tomoyuki [University of Miami, Department of Civil, Architectural, and Environmental Engineering, P.O. Box 248294, McArthur Building, Coral Gables, FL 33124-0630 (United States); Solo-Gabriele, Helena M. [University of Miami, Department of Civil, Architectural, and Environmental Engineering, P.O. Box 248294, McArthur Building, Coral Gables, FL 33124-0630 (United States)]. E-mail: hmsolo@miami.edu; Townsend, Timothy G. [University of Florida, Department of Environmental Engineering Sciences, Gainesville, FL 32611-6450 (United States)

    2007-07-15

    The objective of this study was to evaluate the performance of handheld XRF analyzers on wood that has been treated with a preservative containing arsenic. Experiments were designed to evaluate precision, detection limit, effective depth of analysis, and accuracy of the XRF arsenic readings. Results showed that the precision of the XRF improved with increased sample concentration and longer analysis times. Reported detection limits decreased with longer analysis times to values of less than 1 mg/kg or 18 mg/kg, depending on the model used. The effective depth of analysis was within the top 1.2 cm and 2.0 cm of sample for wood containing natural gradients of chemical preservative and concentration extremes, respectively. XRF results were found to be 1.5-2.3 times higher than measurements from traditional laboratory analysis. Equations can be developed to convert XRF values to results which are consistent with traditional laboratory testing. - Handheld XRF analyzers provided quantitative results for the amount of arsenic within preservative-treated wood.

  6. User certification of hand-held x-ray tube based analytical fluorescent devices in a canadian context

    International Nuclear Information System (INIS)

    Maharaj, H.P.

    2005-01-01

    Safety education aims to reduce personal injury and improve well being. This health promotion principle is applied in the case of hand-held open beam x-ray tube based analytical x-ray devices. Such devices not only are light weight and portable, but also present high radiation exposure levels at the beam exit port and potentially can be used in a variety of industrial applications for determination of material composition. There is much potential for radiation risks to occur with resultant adverse effects if such devices are not used by trained individuals within controlled environments. A level of radiation safety knowledge and understanding of the device design, construction and performance characteristics appear warranted. To reduce radiation risks, user certification at a federal level was introduced in 2004 based on International Standards Organization 20807, since that standard comprises elements commensurate with risk reduction strategies. Within these contexts, a federally certified user is deemed to have acquired a level of safety knowledge and skills to facilitate safe use of the device. Certification, however, does not absolve the holder from obligations of compliance with applicable provincial, territorial or federal laws respecting device operation. The union of federal certification and applicable legislative mandated operational criteria reduces radiation risks overall. (author)

  7. Evaluation of portable Raman spectroscopy and handheld X-ray fluorescence analysis (hXRF) for the direct analysis of glyptics

    Science.gov (United States)

    Lauwers, D.; Candeias, A.; Coccato, A.; Mirao, J.; Moens, L.; Vandenabeele, P.

    2016-03-01

    In archaeometry, the advantages of a combined use of Raman spectroscopy and X-ray fluorescence spectroscopy are extensively discussed for applications such as the analysis of paintings, manuscripts, pottery, etc. Here, we demonstrate for the first time the advantage of using both techniques for analysing glyptics. These engraved gemstones or glass materials were originally used as stamps, to identify the owner, for instance on letters, but also on wine vessels. For this research, a set of 64 glyptics (42 Roman glass specimens and 22 modern ones), belonging to the collection of the museum 'Quinta das Cruzes' in Funchal (Madeira, Portugal), was analysed with portable Raman spectroscopy and handheld X-ray fluorescence (hXRF). These techniques were also used to confirm the gemological identification of these precious objects and can give extra information about the glass composition. Raman spectroscopy identifies the molecular composition as well as on the crystalline phases present. On the other hand, hXRF results show that the antique Roman glass samples are characterised with low Pb and Sn levels and that the modern specimens can be discriminated in two groups: lead-based and non-lead-based ones.

  8. X-ray fluorescence holography

    CERN Document Server

    Hayashi, K; Takahashi, Y

    2003-01-01

    X-ray fluorescence holography (XFH) is a new structural analysis method of determining a 3D atomic arrangement around fluorescing atoms. We developed an XFH apparatus using advanced X-ray techniques and succeeded in obtaining high-quality hologram data. Furthermore, we introduced applications to the structural analysis of a thin film and the environment around dopants and, discussed the quantitative analysis of local lattice distortion. (author)

  9. Elemental Analysis of Bone, Teeth, Horn and Antler in Different Animal Species Using Non-Invasive Handheld X-Ray Fluorescence.

    Science.gov (United States)

    Buddhachat, Kittisak; Klinhom, Sarisa; Siengdee, Puntita; Brown, Janine L; Nomsiri, Raksiri; Kaewmong, Patcharaporn; Thitaram, Chatchote; Mahakkanukrauh, Pasuk; Nganvongpanit, Korakot

    2016-01-01

    Mineralized tissues accumulate elements that play crucial roles in animal health. Although elemental content of bone, blood and teeth of human and some animal species have been characterized, data for many others are lacking, as well as species comparisons. Here we describe the distribution of elements in horn (Bovidae), antler (Cervidae), teeth and bone (humerus) across a number of species determined by handheld X-ray fluorescence (XRF) to better understand differences and potential biological relevance. A difference in elemental profiles between horns and antlers was observed, possibly due to the outer layer of horns being comprised of keratin, whereas antlers are true bone. Species differences in tissue elemental content may be intrinsic, but also related to feeding habits that contribute to mineral accumulation, particularly for toxic heavy metals. One significant finding was a higher level of iron (Fe) in the humerus bone of elephants compared to other species. This may be an adaptation of the hematopoietic system by distributing Fe throughout the bone rather than the marrow, as elephant humerus lacks a marrow cavity. We also conducted discriminant analysis and found XRF was capable of distinguishing samples from different species, with humerus bone being the best source for species discrimination. For example, we found a 79.2% correct prediction and success rate of 80% for classification between human and non-human humerus bone. These findings show that handheld XRF can serve as an effective tool for the biological study of elemental composition in mineralized tissue samples and may have a forensic application.

  10. X-ray fluorescence holography.

    Science.gov (United States)

    Hayashi, Kouichi; Happo, Naohisa; Hosokawa, Shinya; Hu, Wen; Matsushita, Tomohiro

    2012-03-07

    X-ray fluorescence holography (XFH) is a method of atomic resolution holography which utilizes fluorescing atoms as a wave source or a monitor of the interference field within a crystal sample. It provides three-dimensional atomic images around a specified element and has a range of up to a few nm in real space. Because of this feature, XFH is expected to be used for medium-range local structural analysis, which cannot be performed by x-ray diffraction or x-ray absorption fine structure analysis. In this article, we explain the theory of XFH including solutions to the twin-image problem, an advanced measuring system, and data processing for the reconstruction of atomic images. Then, we briefly introduce our recent applications of this technique to the analysis of local lattice distortions in mixed crystals and nanometer-size clusters appearing in the low-temperature phase of a shape-memory alloy.

  11. X-ray fluorescence holography

    International Nuclear Information System (INIS)

    Hayashi, Kouichi; Happo, Naohisa; Hosokawa, Shinya; Hu Wen; Matsushita, Tomohiro

    2012-01-01

    X-ray fluorescence holography (XFH) is a method of atomic resolution holography which utilizes fluorescing atoms as a wave source or a monitor of the interference field within a crystal sample. It provides three-dimensional atomic images around a specified element and has a range of up to a few nm in real space. Because of this feature, XFH is expected to be used for medium-range local structural analysis, which cannot be performed by x-ray diffraction or x-ray absorption fine structure analysis. In this article, we explain the theory of XFH including solutions to the twin-image problem, an advanced measuring system, and data processing for the reconstruction of atomic images. Then, we briefly introduce our recent applications of this technique to the analysis of local lattice distortions in mixed crystals and nanometer-size clusters appearing in the low-temperature phase of a shape-memory alloy. (topical review)

  12. X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Vries, J.L. de.

    1976-01-01

    The seventh edition of Philips' Review of Literature on x-ray fluorescence spectrometry starts with a list of conference proceedings on the subject, organised by the Philips organisation at regular intervals in various European countries. It is followed by a list of bulletins. The bibliography is subdivided according to spectra, equipment, applications and absorption analysis

  13. X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Ray, N.B.

    1977-01-01

    The principle, instrument and procedure of X-ray fluorescence spectrometry are described. It is a rapid, simple and sensitive method for the trace analysis of elements from sodium to uranium in powder, liquid or metal samples. (M.G.B.)

  14. Elemental Analysis of Bone, Teeth, Horn and Antler in Different Animal Species Using Non-Invasive Handheld X-Ray Fluorescence

    Science.gov (United States)

    Buddhachat, Kittisak; Klinhom, Sarisa; Siengdee, Puntita; Brown, Janine L.; Nomsiri, Raksiri; Kaewmong, Patcharaporn; Thitaram, Chatchote; Mahakkanukrauh, Pasuk; Nganvongpanit, Korakot

    2016-01-01

    Mineralized tissues accumulate elements that play crucial roles in animal health. Although elemental content of bone, blood and teeth of human and some animal species have been characterized, data for many others are lacking, as well as species comparisons. Here we describe the distribution of elements in horn (Bovidae), antler (Cervidae), teeth and bone (humerus) across a number of species determined by handheld X-ray fluorescence (XRF) to better understand differences and potential biological relevance. A difference in elemental profiles between horns and antlers was observed, possibly due to the outer layer of horns being comprised of keratin, whereas antlers are true bone. Species differences in tissue elemental content may be intrinsic, but also related to feeding habits that contribute to mineral accumulation, particularly for toxic heavy metals. One significant finding was a higher level of iron (Fe) in the humerus bone of elephants compared to other species. This may be an adaptation of the hematopoietic system by distributing Fe throughout the bone rather than the marrow, as elephant humerus lacks a marrow cavity. We also conducted discriminant analysis and found XRF was capable of distinguishing samples from different species, with humerus bone being the best source for species discrimination. For example, we found a 79.2% correct prediction and success rate of 80% for classification between human and non-human humerus bone. These findings show that handheld XRF can serve as an effective tool for the biological study of elemental composition in mineralized tissue samples and may have a forensic application. PMID:27196603

  15. Elemental Analysis of Bone, Teeth, Horn and Antler in Different Animal Species Using Non-Invasive Handheld X-Ray Fluorescence.

    Directory of Open Access Journals (Sweden)

    Kittisak Buddhachat

    Full Text Available Mineralized tissues accumulate elements that play crucial roles in animal health. Although elemental content of bone, blood and teeth of human and some animal species have been characterized, data for many others are lacking, as well as species comparisons. Here we describe the distribution of elements in horn (Bovidae, antler (Cervidae, teeth and bone (humerus across a number of species determined by handheld X-ray fluorescence (XRF to better understand differences and potential biological relevance. A difference in elemental profiles between horns and antlers was observed, possibly due to the outer layer of horns being comprised of keratin, whereas antlers are true bone. Species differences in tissue elemental content may be intrinsic, but also related to feeding habits that contribute to mineral accumulation, particularly for toxic heavy metals. One significant finding was a higher level of iron (Fe in the humerus bone of elephants compared to other species. This may be an adaptation of the hematopoietic system by distributing Fe throughout the bone rather than the marrow, as elephant humerus lacks a marrow cavity. We also conducted discriminant analysis and found XRF was capable of distinguishing samples from different species, with humerus bone being the best source for species discrimination. For example, we found a 79.2% correct prediction and success rate of 80% for classification between human and non-human humerus bone. These findings show that handheld XRF can serve as an effective tool for the biological study of elemental composition in mineralized tissue samples and may have a forensic application.

  16. X-ray emission spectroscopy. X-ray fluorescence

    International Nuclear Information System (INIS)

    Despujols, J.

    1992-01-01

    Principles of X-ray emission spectrometry are first recalled, then wave-length dispersive and energy dispersive X-ray fluorescence spectrometer are described. They are essentially designed for qualitative and quantitative analysis of elements (Z>10). Sample preparation, calibration, corrections, interferences, accuracy are reviewed. Examples of use in different industries are given. (71 refs.)

  17. Radionuclide X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Cechak, T.

    1994-01-01

    The author's achievements in the title field are summarized and discussed. The following topics are dealt with: (i) principles of radionuclide X-ray fluorescence analysis; (ii) mathematical methods in X-ray fluorescence analysis; (iii) Ross differential filters; (iv) application of radionuclide X-ray fluorescence analysis in the coal industry (with emphasis on the determination of the ash content, sulfur content, and arsenic content of coal); and (v) evaluation of the X-ray fluorescence analyzer from the radiological safety point of view. (P.A.)

  18. In situ determination of K, Ca, S and Si in fresh sugar cane leaves by handheld Energy Dispersive X-Ray Fluorescence Spectrometry

    Energy Technology Data Exchange (ETDEWEB)

    Guerra, Marcelo B.B.; Adame, Andressa; Almeida, Eduardo de; Brasil, Marcos A.S.; Krug, Francisco J., E-mail: fjkrug@cena.usp.br [Centro de Energia Nuclear na Agricultura, Piracicaba, SP (Brazil); Schaefer, Carlos E.G.R. [Departamento de Solos, Universidade Federal de Viçosa, MG (Brazil)

    2018-05-01

    A portable energy dispersive X-ray fluorescence spectrometer was evaluated in the in situ analysis of fresh sugar cane leaves for real time plant nutrition diagnosis. Fresh leaf fragments (n = 10 sugar cane varieties; 20 fragments per leaf; 2 measurement sites per fragment) were irradiated and the averaged data from X-ray characteristic emission lines intensities (for K, Ca, S and Si Kα lines) were in close agreement with mass fraction data obtained by a validated comparative method. The linear correlation coefficients (r) ranged from 0.9575 for Ca to 0.9851 for Si. The obtained limits of detection were at least two-fold lower than the critical nutrient levels. Manganese can also be properly determined, but validation still requires more robust calibration models. The proposed method is a straightforward approach towards the fast evaluation of the nutritional profile of plants avoiding time-consuming steps, which involve drying, grinding, weighing, and acid digestion. (author)

  19. Automated x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    O'Connell, A.M.

    1977-01-01

    A fully automated x-ray fluorescence analytical system is described. The hardware is based on a Philips PW1220 sequential x-ray spectrometer. Software for on-line analysis of a wide range of sample types has been developed for the Hewlett-Packard 9810A programmable calculator. Routines to test the system hardware are also described. (Author)

  20. X-ray fluorescence in geology

    International Nuclear Information System (INIS)

    Dutra, C.V.; Gomes, C.B.

    1990-01-01

    This work is about the X-ray fluorescence aplication in geology. It's showing the X-ray origin and excitation. About the instrumentation this work shows the following: X-ray tubes, colimators, analysers crystals, detectors, amplifiers, pulse height selector, and others electronic components. By X-ray fluorescente are done quantitative and qualitative geological analysis and this work shows this analysis and its detection limits. The problems determination is the example. In this work was done yet the comparative analysis of the various instrumental methods in geochemistry. (C.G.) [pt

  1. X-ray fluorescence analyzer arrangement

    International Nuclear Information System (INIS)

    Vatai, Endre; Ando, Laszlo; Gal, Janos.

    1981-01-01

    An x-ray fluorescence analyzer for the quantitative determination of one or more elements of complex samples is reported. The novelties of the invention are the excitation of the samples by x-rays or γ-radiation, the application of a balanced filter pair as energy selector, and the measurement of the current or ion charge of ionization detectors used as sensors. Due to the increased sensitivity and accuracy, the novel design can extend the application fields of x-ray fluorescence analyzers. (A.L.)

  2. Preliminary Study to Test the Feasibility of Sex Identification of Human (Homo sapiens) Bones Based on Differences in Elemental Profiles Determined by Handheld X-ray Fluorescence.

    Science.gov (United States)

    Nganvongpanit, Korakot; Buddhachat, Kittisak; Brown, Janine L; Klinhom, Sarisa; Pitakarnnop, Tanita; Mahakkanukrauh, Pasuk

    2016-09-01

    Sex assignment of human remains is a crucial step in forensic anthropological studies. The aim of this study was to examine elemental differences between male and female bones using X-ray fluorescence (XRF) and determine if elemental profiling could be used for sex discrimination. Cranium, humerus, and os coxae of 60 skeletons (30 male, 30 female) from the Chiang Mai University Skeletal Collection were scanned by XRF and differences in elemental profiles between male and female bones determined using discriminant analysis. In the cranium, three elements (S, Ca, Pb) were significantly higher in males and five elements (Si, Mn, Fe, Zn, Ag) plus light elements (atomic number lower than 12) were higher in females. In humerus and os coxae, nine elements were significantly higher in male and one element was higher in female samples. The accuracy rate for sex estimation was 60, 63, and 61 % for cranium, humerus, and os coxae, respectively, and 67 % when data for all three bones were combined. We conclude that there are sex differences in bone elemental profiles; however, the accuracy of XRF analyses for discriminating between male and female samples was low compared to standard morphometric and molecular methods. XRF could be used on small samples that cannot be sexed by traditional morphological methods, but more work is needed to increase the power of this technique for gender assignment.

  3. The Use of Handheld X-Ray Fluorescence (XRF) Technology in Unraveling the Eruptive History of the San Francisco Volcanic Field, Arizona

    Science.gov (United States)

    Young, Kelsey E.; Evans, C. A.; Hodges, K. V.

    2012-01-01

    While traditional geologic mapping includes the examination of structural relationships between rock units in the field, more advanced technology now enables us to simultaneously collect and combine analytical datasets with field observations. Information about tectonomagmatic processes can be gleaned from these combined data products. Historically, construction of multi-layered field maps that include sample data has been accomplished serially (first map and collect samples, analyze samples, combine data, and finally, readjust maps and conclusions about geologic history based on combined data sets). New instruments that can be used in the field, such as a handheld xray fluorescence (XRF) unit, are now available. Targeted use of such instruments enables geologists to collect preliminary geochemical data while in the field so that they can optimize scientific data return from each field traverse. Our study tests the application of this technology and projects the benefits gained by real-time geochemical data in the field. The integrated data set produces a richer geologic map and facilitates a stronger contextual picture for field geologists when collecting field observations and samples for future laboratory work. Real-time geochemical data on samples also provide valuable insight regarding sampling decisions by the field geologist

  4. Capacity of X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Wobrauschek, P.; Kregsamer, P.

    1997-01-01

    X-Ray fluorescence analysis (XRF) is a powerful analytical tool for the qualitative and quantitative determination of chemical elements in a sample. Two different detection principles are accepted widely: wavelength dispersive and energy dispersive. Various sources for XRF are discussed: X-ray tubes, accelerators for particle induced XRF, radioactive isotopes, and the use of synchrotron radiation. Applications include environmental, technical, medical, fine art, and forensic studies. Due to the demands of research and application special techniques like total reflection XRF (TXRF) were developed with ultimately achievable detection limits in the femtogram region. The elements detectable by XRF range from Be to U. (author)

  5. Escape probabilities for fluorescent x-rays

    International Nuclear Information System (INIS)

    Dance, D.R.; Day, G.J.

    1985-01-01

    Computation of the energy absorption efficiency of an x-ray photon detector involves consideration of the histories of the secondary particles produced in any initial or secondary interaction which may occur within the detector. In particular, the K or higher shell fluorescent x-rays which may be emitted following a photoelectric interaction can carry away a large fraction of the energy of the incident photon, especially if this energy is just above an absorption edge. The effects of such photons cannot be ignored and a correction term, depending upon the probability that the fluorescent x-rays will escape from the detector, must be applied to the energy absorption efficiency. For detectors such as x-ray intensifying screens, it has been usual to calculate this probability by numerical integration. In this note analytic expressions are derived for the escape probability of fluorescent photons from planar detectors in terms of exponential integral functions. Rational approximations for these functions are readily available and these analytic expressions therefore facilitate the computation of photon absorption efficiencies. A table is presented which should obviate the need for calculating the escape probability for most cases of interest. (author)

  6. Development of confocal micro X-ray fluorescence instrument using two X-ray beams

    International Nuclear Information System (INIS)

    Tsuji, Kouichi; Nakano, Kazuhiko; Ding Xunliang

    2007-01-01

    A new confocal micro X-ray fluorescence instrument was developed. This instrument has two independent micro X-ray tubes with Mo targets. A full polycapillary X-ray lens was attached to each X-ray tube. Another half polycapillary lens was attached to a silicon drift X-ray detector (SDD). The focal spots of the three lenses were adjusted to a common position. The effects of the excitation of two X-ray beams were investigated. The instrument enabled highly sensitive three-dimensional X-ray fluorescence analysis. We confirmed that the X-ray fluorescence intensity from the sample increased by applying the two independent X-ray tubes in confocal configuration. Elemental depth profiling of black wheat was demonstrated with the result that each element in the surface coat of a wheat grain showed unique distribution

  7. Millianalyser by x-ray fluorescence

    International Nuclear Information System (INIS)

    Kawamoto, A.; Hirao, O.; Kashiwakura, J.; Gohshi, Y.

    1976-01-01

    Research on the possibility of mm-size nondestructive analysis was carried out by the fluorescent x-ray method. With 0.2 mm pin-hole slit, source x-rays from a Cu target diffraction tube were collimated to a spot smaller than 1 mm phi at a slide stage placed about 5 cm distant from the pin-hole slit. Resultant x-rays from a sample placed on the slide stage, which is excited by the collimated x-ray, were detected with a head-on-type 6 mm SSD, placed so that its 12.5 micron Be window was about 5 cm beneath the stage. X-ray intensities sufficient for analysis (500 to 5000 CPS) could be obtained for various metallic samples with up to 40 kV-10 mA excitation. This instrument proved to be useful for mm-size qualitative analysis in measurements of tiny samples. Furthermore, the possibility of distribution analysis is expected based on the result of an investigation on c.a. 0.1 percent Cr in LiNbO 3 , where the ratios of Cr-Kα intensity to scattered Cu-Kα intensity varied between 0.094 and 0.19, with deviations of less than 7.5 percent at five successive points located at 2 mm intervals along the direction of growth

  8. Synchrotron x-ray microbeam characteristics for x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Iida, Atsuo; Noma, Takashi

    1995-01-01

    X-ray fluorescence analysis using a synchrotron x-ray microprobe has become an indispensable technique for non-destructive micro-analysis. One of the most important parameters that characterize the x-ray microbeam system for x-ray fluorescence analysis is the beam size. For practical analysis, however, the photon flux, the energy resolution and the available energy range are also crucial. Three types of x-ray microbeam systems, including monochromatic and continuum excitation systems, were compared with reference to the sensitivity, the minimum detection limit and the applicability to various types of x-ray spectroscopic analysis. 16 refs., 5 figs

  9. X-ray fluorescence imaging with polycapillary X-ray optics

    International Nuclear Information System (INIS)

    Yonehara, Tasuku; Yamaguchi, Makoto; Tsuji, Kouichi

    2010-01-01

    X-ray fluorescence spectrometry imaging is a powerful tool to provide information about the chemical composition and elemental distribution of a specimen. X-ray fluorescence spectrometry images were conventionally obtained by using a μ-X-ray fluorescence spectrometry spectrometer, which requires scanning a sample. Faster X-ray fluorescence spectrometry imaging would be achieved by eliminating the process of sample scanning. Thus, we developed an X-ray fluorescence spectrometry imaging instrument without sample scanning by using polycapillary X-ray optics, which had energy filter characteristics caused by the energy dependence of the total reflection phenomenon. In the present paper, we show that two independent straight polycapillary X-ray optics could be used as an energy filter of X-rays for X-ray fluorescence. Only low energy X-rays were detected when the angle between the two optical axes was increased slightly. Energy-selective X-ray fluorescence spectrometry images with projection mode were taken by using an X-ray CCD camera equipped with two polycapillary optics. It was shown that Fe Kα (6.40 keV) and Cu Kα (8.04 keV) could be discriminated for Fe and Cu foils.

  10. X-ray microtome by fluorescence tomography

    CERN Document Server

    Simionovici, A S; Guenzler, F; Schrör, C; Snigirev, A; Snigireva, I; Tümmler, J; Weitkamp, T

    2001-01-01

    The X-ray fluorescence microtomography method is presented, which is capable of virtually slicing samples to obtain cross-sections of their inner structure. High precision experimental results of fluo-tomography in 'pencil-beam' geometry with up to 1.2 mu m resolution are described. Image reconstructions are based on either a simplified algebraic reconstruction method (ART) or the filtered back-projection method (FBP). Phantoms of inhomogeneous test objects as well as biological samples are successfully analyzed.

  11. Total reflection X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Michaelis, W.; Prange, A.

    1987-01-01

    In the past few years, total reflection X-ray flourescence analysis (TXRF) has found an increasing number of assignments and applications. Experience of trace element analysis using TXRF and examples of applications are already widespread. Therefore, users of TXRF had the opportunity of an intensive exchange of their experience at the 1st workshop on total reflection X-ray fluorescence analysis which took place on May 27th and 28th 1986 at the GKSS Research Centre at Geesthacht. In a series of lectures and discussions dealing with the analytical principle itself, sample preparation techniques and applications as well as comuter programs for spectrum evaluation, the present state of development and the range of applications were outlined. 3 studies out of a total of 14 were included separately in the INIS and ENERGY databases. With 61 figs., 12 tabs [de

  12. Silicon lithium detector for x ray fluorescence

    International Nuclear Information System (INIS)

    Rodriguez Cabal, A. E.; Diaz Garcia, A.; Noriega Scull, C.; Martinez Munoz, O.; Diaz Cepeda, R.

    1997-01-01

    The Silicon Lithium detector is the system for the detection of nuclear radiation. It transforms the charge that was produced inside of Silicon material as a result of the incidence of particles and X rays, in voltage pulses at the output of the preamplifier. In this work was made the adjustment of the technological process of manufacture of the detector. Also was made the design and construction of the cryostat and preamplifier and then the validation of the system in a Cuban Dewar. The system, which was made for the first time in our country, has an energy resolution of 185 eV for the Fe-55 source (E=5.9 KeV), which has permitted its implementation in energy dispersive X ray fluorescence. (author) [es

  13. Submicron, soft x-ray fluorescence imaging

    International Nuclear Information System (INIS)

    La Fontaine, B.; MacDowell, A.A.; Tan, Z.; White, D.L.; Taylor, G.N.; Wood, O.R. II; Bjorkholm, J.E.; Tennant, D.M.; Hulbert, S.L.

    1995-01-01

    Submicron fluorescence imaging of soft x-ray aerial images, using a high resolution fluorescent crystal is reported. Features as small as 0.1 μm were observed using a commercially available single-crystal phosphor, STI-F10G (Star Tech Instruments Inc. P. O. Box 2536, Danbury, CT 06813-2536), excited with 139 A light. Its quantum efficiency was estimated to be 5--10 times that of sodium salicylate and to be constant over a broad spectral range from 30 to 400 A. A comparison with a terbium-activated yttrium orthosilicate fluorescent crystal is also presented. Several applications, such as the characterization of the aerial images produced by deep ultraviolet or extreme ultraviolet lithographic exposure tools, are envisaged

  14. X-ray fluorescence method for trace analysis and imaging

    International Nuclear Information System (INIS)

    Hayakawa, Shinjiro

    2000-01-01

    X-ray fluorescence analysis has a long history as conventional bulk elemental analysis with medium sensitivity. However, with the use of synchrotron radiation x-ray fluorescence method has become a unique analytical technique which can provide tace elemental information with the spatial resolution. To obtain quantitative information of trace elemental distribution by using the x-ray fluorescence method, theoretical description of x-ray fluorescence yield is described. Moreover, methods and instruments for trace characterization with a scanning x-ray microprobe are described. (author)

  15. In vivo X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Ahlgren, L.

    1980-02-01

    Measurements on five occupationally exposed persons have shown that it is possible to use X-ray fluorescence analysis for in vivo measurements of lead in the skeleton. The technique for calibrating in vivo X-ray fluorescence measurements of lead in bone tissue has been studied in detail and a two-component phantom simulating the bone and the soft tissue parts of the finger constructed. The technique has been used for in vivo measurements on 22 occupationally exposed persons. The minimum detectable concentration of lead in fingerbones was found to be around 20 μg x g -1 . The lead concentrations in their skeletons and blood were compared: the correlation was poor. The variations in lead concentrations in the skeleton have been studied in occupationally exposed persons and in samples from archaeological skeletons. The sensitivity and the minimum detectable concentration of cadmium in the kidney cortex in in vivo measurements has been studied by measurements on kidney models. The minimum detectable concentration was 20 μg x g -1 at a skin-kidney distance of 30 mm and 40 μg x g -1 at 40 mm. Five persons occupationally exposed were studied. (Author)

  16. X-ray Fluorescence Holography: Principles, Apparatus, and Applications

    Science.gov (United States)

    Hayashi, Kouichi; Korecki, Pawel

    2018-06-01

    X-ray fluorescence holography (XFH) is an atomic structure determination technique that combines the capabilities of X-ray diffraction and X-ray fluorescence spectroscopy. It provides a unique means of gaining fully three-dimensional information about the local atomic structure and lattice site positions of selected elements inside compound samples. In this work, we discuss experimental and theoretical aspects that are essential for the efficient recording and analysis of X-ray fluorescence holograms and review the most recent advances in XFH. We describe experiments performed with brilliant synchrotron radiation as well as with tabletop setups that employ conventional X-ray tubes.

  17. Leakage and scattered radiation from hand-held dental x-ray unit

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Eun Kyung [Dankook Univ. School of Dentistry, Seoul (Korea, Republic of)

    2007-06-15

    To compare the leakage and scattered radiation from hand-held dental X-ray unit with radiation from fixed dental X-ray unit. For evaluation we used one hand-held dental X-ray unit and Oramatic 558 (Trophy Radiologie, France), a fixed dental X-ray unit. Doses were measured with Unfors Multi-O-Meter 512L at the right and left hand levels of X-ray tube head part for the scattered and leakage radiation when human skull DXTTR {iota}{iota}{iota} was exposed to both dental X-ray units. And for the leakage radiation only, doses were measured at the immediately right, left, superior and posterior side of the tube head part when air was exposed. Exposure parameters of hand-held dental X-ray unit were 70 kVp, 3 mA , 0.1 second, and of fixed X-ray unit 70 kVp, 8 mA, 0.45 second. The mean dose at the hand level when human skull DXTTR {iota}{iota}{iota} was exposed with portable X-ray unit 6.39 {mu}Gy, and the mean dose with fixed X-ray unit 3.03 {mu}Gy (p<0.001). The mean dose at the immediate side of the tube head part when air was exposed with portable X-ray unit was 2.97 {mu}Gy and with fixed X-ray unit the mean dose was 0.68 {mu}Gy (p<0.01). The leakage and scattered radiation from hand-held dental radiography was greater than from fixed dental radiography.

  18. Leakage and scattered radiation from hand-held dental x-ray unit

    International Nuclear Information System (INIS)

    Kim, Eun Kyung

    2007-01-01

    To compare the leakage and scattered radiation from hand-held dental X-ray unit with radiation from fixed dental X-ray unit. For evaluation we used one hand-held dental X-ray unit and Oramatic 558 (Trophy Radiologie, France), a fixed dental X-ray unit. Doses were measured with Unfors Multi-O-Meter 512L at the right and left hand levels of X-ray tube head part for the scattered and leakage radiation when human skull DXTTR ΙΙΙ was exposed to both dental X-ray units. And for the leakage radiation only, doses were measured at the immediately right, left, superior and posterior side of the tube head part when air was exposed. Exposure parameters of hand-held dental X-ray unit were 70 kVp, 3 mA , 0.1 second, and of fixed X-ray unit 70 kVp, 8 mA, 0.45 second. The mean dose at the hand level when human skull DXTTR ΙΙΙ was exposed with portable X-ray unit 6.39 μGy, and the mean dose with fixed X-ray unit 3.03 μGy (p<0.001). The mean dose at the immediate side of the tube head part when air was exposed with portable X-ray unit was 2.97 μGy and with fixed X-ray unit the mean dose was 0.68 μGy (p<0.01). The leakage and scattered radiation from hand-held dental radiography was greater than from fixed dental radiography

  19. Advanced in X-ray fluorescence holography

    CERN Document Server

    Hayashi, K

    2002-01-01

    X-ray fluorescence holography (XFH) can resolve 'phase problem' in crystal diffraction and therefore it provides 3D atomic images around specific elements. Since first demonstration of the XFH in 1996, view of atoms has been improved rapidly with the refinement of the hologram data collection method. The present performance of the XFH makes it possible to apply to impurity, thin film and quasicrystal, and opens a way to practical tool for determination of local structure. In this paper, theory including solutions for twin image problem, advanced experimental systems and application to Si sub 0 sub . sub 9 sub 9 sub 9 Ge sub 0 sub . sub 0 sub 0 sub 1 are discussed. (author)

  20. Review of X-ray Tomography and X-ray Fluorescence Spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Shear, Trevor A. [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-03-16

    This literature review will focus on both laboratory and synchrotron based X-ray tomography of materials and highlight the inner workings of these instruments. X-ray fluorescence spectroscopy will also be reviewed and applications of the tandem use of these techniques will be explored. The real world application of these techniques during the internship will also be discussed.

  1. Position statement on use of hand-held portable dental X-ray equipment

    International Nuclear Information System (INIS)

    2014-06-01

    The position statement focuses on justification in the medical field, in particular on the use of hand-held portable dental x-ray equipment. It supplements another HERCA position paper, providing a general overview of the use of all hand-held portable X-ray equipment. Key Messages: - HERCA finds that the use of hand-held portable X-ray devices should be discouraged except in special circumstances. - As a general rule, these devices should only be used in scenarios where an intraoral radiograph is deemed necessary for a patient and the use of a fixed or semi-mobile x-ray unit is impractical, e.g.: - nursing homes, residential care facilities or homes for persons with disabilities; - forensic odontology, - military operations abroad without dental facilities

  2. Fluorescent scanning x-ray tomography with synchrotron radiation

    Science.gov (United States)

    Takeda, Tohoru; Maeda, Toshikazu; Yuasa, Tetsuya; Akatsuka, Takao; Ito, Tatsuo; Kishi, Kenichi; Wu, Jin; Kazama, Masahiro; Hyodo, Kazuyuki; Itai, Yuji

    1995-02-01

    Fluorescent scanning (FS) x-ray tomography was developed to detect nonradioactive tracer materials (iodine and gadolinium) in a living object. FS x-ray tomography consists of a silicon (111) channel cut monochromator, an x-ray shutter, an x-ray slit system and a collimator for detection, a scanning table for the target organ, and an x-ray detector with pure germanium. The minimal detectable dose of iodine in this experiment was 100 ng in a volume of 2 mm3 and a linear relationship was shown between the photon counts of a fluorescent x ray and the concentration of iodine contrast material. A FS x-ray tomographic image was clearly obtained with a phantom.

  3. A library for X-ray-matter interaction cross sections for X-ray fluorescence applications

    Energy Technology Data Exchange (ETDEWEB)

    Brunetti, A. [Istituto di Matematica e Fisica, Universita di Sassari, via Vienna 2, 07100 Sassari (Italy) and INFN, Sezione di Cagliari (Italy)]. E-mail: brunetti@uniss.it; Sanchez del Rio, M. [European Synchrotron Radiation Facility, 6 rue Jules Horowitz, 38043 Grenoble Cedex (France); Golosio, B. [INFN, Sezione di Cagliari (Italy); European Synchrotron Radiation Facility, 6 rue Jules Horowitz, 38043 Grenoble Cedex (France); Simionovici, A. [European Synchrotron Radiation Facility, 6 rue Jules Horowitz, 38043 Grenoble Cedex (France); Laboratoire de Sciences de la Terre, Ecole Normale Superieure, Lyon, F-69364 (France); Somogyi, A. [European Synchrotron Radiation Facility, 6 rue Jules Horowitz, 38043 Grenoble Cedex (France)

    2004-10-08

    Quantitative estimate of elemental composition by spectroscopic and imaging techniques using X-ray fluorescence requires the availability of accurate data of X-ray interaction with matter. Although a wide number of computer codes and data sets are reported in literature, none of them is presented in the form of freely available library functions which can be easily included in software applications for X-ray fluorescence. This work presents a compilation of data sets from different published works and an xraylib interface in the form of callable functions. Although the target applications are on X-ray fluorescence, cross sections of interactions like photoionization, coherent scattering and Compton scattering, as well as form factors and anomalous scattering functions, are also available.

  4. Advanced of X-ray fluorescence logging technique in China

    International Nuclear Information System (INIS)

    Zhou Sichun; Ge Liangquan; Lai Wanchang; Yang Qiang

    2010-01-01

    The paper discuses principle of X-ray fluorescence logging, and introduces advanced of X-ray fluorescence logging technique in China. By 2009, third generation XRF logging instrument has been developed in China, and good logging result has been obtained in Lala copper mine. (authors)

  5. Portable X-ray fluorescence analyzer of high sensitivity using X-ray tube excitation

    International Nuclear Information System (INIS)

    Vatai, E.; Ando, L.

    1982-01-01

    A review of the three main methods of X-ray fluorescence analysis and their problems is given. The attainable accuracy and effectiveness of each method are discussed. The main properties of portable X-ray analyzers required by the industry are described. The results and experiences of R and D activities in ATOMKI (Debrecen, Hungary) for developing portable X-ray analyzers are presented. The only way for increasing the accuracy and decreasing the measuring time is the application of X-ray tube excitation instead of radioactive sources. The new ATOMKI equipment presently under construction and patenting uses X-ray tube excitation; it will increase the accuracy of concentration determination by one order of magnitude. (D.Gy.)

  6. X-ray Microprobe for Fluorescence and Diffraction Analysis

    International Nuclear Information System (INIS)

    Ice, G.E.

    2005-01-01

    X-ray diffraction (see unit 1.1) and x-ray excited fluorescence analysis are powerful techniques for the nondestructive measurement of crystal structure and chemical composition. X-ray fluorescence analysis is inherently nondestructive with orders of magnitude lower power deposited for the same detectable limit as with fluorescence excited by charged particle probes (Sparks, 1980). X-ray diffraction analysis is sensitive to crystal structure with orders-of-magnitude greater sensitivity to crystallographic strain than electron probes (Rebonato, et al. 1989). When a small-area x-ray microbeam is used as the probe, chemical composition (Z>14), crystal structure, crystalline texture, and crystalline strain distributions can be determined. These distributions can be studied both at the surface of the sample and deep within the sample (Fig. 1). Current state-of-the-art can achieve an ∼1 mm-D x-ray microprobe and an ∼0.1 mm-D x-ray microprobe has been demonstrated (Bilderback, et al., 1994). Despite their great chemical and crystallographic sensitivities, x-ray microprobe techniques have until recently been restricted by inefficient x-ray focusing optics and weak x-ray sources; x-ray microbeam analysis was largely superseded by electron techniques in the 50's. However, interest in x-ray microprobe techniques has now been revived (Howells, et al., 1983; Ice and Sparks, 1984; Chevallier, et al., 1997; Riekel 1992; Thompson, el al., 1992; and Making and Using... 1997) by the development of efficient x-ray focusing optics and ultra-high intensity synchrotron x-ray sources (Buras and Tazzari, 1984; Shenoy, et al., 1988). These advances have increased the achievable microbeam flux by more than 11 orders of magnitude (Fig. 2) (Ice, 1997); the flux in a tunable 1 mm-D beam on a 'so called' 3rd-generation synchrotron source such as the APS can exceed the flux in a fixed-energy mm2 beam on a conventional source. These advances make x-ray microfluorescence and x-ray

  7. Quantitative analysis with energy dispersive X-ray fluorescence analyser

    International Nuclear Information System (INIS)

    Kataria, S.K.; Kapoor, S.S.; Lal, M.; Rao, B.V.N.

    1977-01-01

    Quantitative analysis of samples using radioisotope excited energy dispersive x-ray fluorescence system is described. The complete set-up is built around a locally made Si(Li) detector x-ray spectrometer with an energy resolution of 220 eV at 5.94 KeV. The photopeaks observed in the x-ray fluorescence spectra are fitted with a Gaussian function and the intensities of the characteristic x-ray lines are extracted, which in turn are used for calculating the elemental concentrations. The results for a few typical cases are presented. (author)

  8. Preparation of specimens for analysis by: X-ray diffraction and X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Banos L, L.

    2004-01-01

    Specimen preparation is one of the most important requirements in the analysis of samples by X-ray Diffraction and X-ray Fluorescence. This statement is especially true for samples containing different types of materials. There are many forms of specimen suitable for X-ray analysis and the type of the sample as received will generally determine the method of pretreatment. It is convenient to refer to the material received for analysis as the sample, and that, which is actually analyzed as the specimen. The powder Diffraction method assumes that the particles in the specimen are ideally random orientation and that there are enough crystallites in the specimen to achieve a representative intensity distribution for these crystallites. X ray Fluorescence is essentially a comparative method of analysis, it is vital that all standards and unknowns be presented to the spectrometer in a reproducible and identical manner. (Author) 3 refs., 6 figs

  9. Milli X-Ray Fluorescence Spectrometer

    Data.gov (United States)

    Federal Laboratory Consortium — The Eagle III Micro XRF unit is similar to a traditional XRF unit, with the primary difference being that the X-rays are focused by a polycapillary optic into a spot...

  10. Use of X-Ray Fluorescence Spectrometry to Determine Trace ...

    African Journals Online (AJOL)

    This paper deals with application of X-ray fluorescence spectrometry for the detection of trace elements in graphic. An X-ray spectrometer was constructed and used to carry out measurements on graphite spheres impregnated with different chemical elements. The intensities of the lines of these trace elements, as function of ...

  11. Synchrotron x-ray fluorescence and extended x-ray absorption fine structure analysis

    International Nuclear Information System (INIS)

    Chen, J.R.; Gordon, B.M.; Hanson, A.L.; Jones, K.W.; Kraner, H.W.; Chao, E.C.T.; Minkin, J.A.

    1984-01-01

    The advent of dedicated synchrotron radiation sources has led to a significant increase in activity in many areas of science dealing with the interaction of x-rays with matter. Synchrotron radiation provides intense, linearly polarized, naturally collimated, continuously tunable photon beams, which are used to determine not only the elemental composition of a complex, polyatomic, dilute material but also the chemical form of the elements with improved accuracy. Examples of the application of synchrotron radiation include experiments in synchrotron x-ray fluorescence (SXRF) analysis and extended x-ray absorption fine structure (EXAFS) analysis. New synchrotron radiation x-ray microprobes for elemental analysis in the parts per billion range are under construction at several laboratories. 76 references, 24 figures

  12. Modern X-ray spectroscopy 3. X-ray fluorescence holography

    International Nuclear Information System (INIS)

    Hayashi, Kouichi

    2008-01-01

    X-ray fluorescence holography (XFH) provides three dimensional atomic images around specified elements. The XFH uses atoms as a wave source or monitor of interference field within a crystal sample, and therefore it can record both intensity and phase of scattered X-rays. Its current performance makes it possible to apply to ultra thin film, impurity and quasicrystal. In this article, I show the theory including solutions for twin image problem, advanced measuring system, data processing for reconstruction of the atomic images and for obtaining accurate atomic positions, applications using resonant X-ray scattering and X-ray excited optical luminescence, and an example of XFH result on the local structure around copper in silicon steal. (author)

  13. Energy dispersive X-Ray fluorescence spectrometric study of ...

    African Journals Online (AJOL)

    Energy dispersive X-Ray fluorescence spectrometric study of compositional differences in trace elements in dried Moringa oleifera leaves grown in two different agro-ecological locations in Ebonyi State, Nigeria.

  14. X-ray fluorescence analysis of Fe - Ni - Mo systems

    International Nuclear Information System (INIS)

    Belyaev, E.E.; Ershov, A.V.; Mashin, A.I.; Mashin, N.I.; Rudnevskij, N.K.

    1998-01-01

    Procedures for the X-ray fluorescence determination of the composition and thickness of Fe - Ni - Mo thin films and the concentration of elements in thick films of the Fe - Ni - Mo alloy are developed [ru

  15. System for Gamma an X rays fluorescence spectrometric

    International Nuclear Information System (INIS)

    Alonso Abad, D.; Arista Romeu, E.; Bolanos Perez, L. and others

    1997-01-01

    A system for spectrometry of gamma or fluorescence X rays is presented. It sis composed by a Si(Li) semiconductors detector, a charge sensitive preamplifier, a high voltage power supply, a spectrometric amplifier and a monolithic 1024 channels multichannel analyzers or an IBM compatible 4096 channels add - on- card multichannel analyzer. The system can be configured as a 1024 or 4096 channels gamma or fluorescent X rays spectrometer

  16. Fluorescent intensifying screens: contribution of secondary X-rays

    International Nuclear Information System (INIS)

    Barroso, R.C.; Goncalves, O.D.; Eichler, J.; Lopes, R.T.; Cardoso, S.C.

    1996-01-01

    The counting rate and angular distribution of secondary X-rays produced by fluorescent intensifying screens are studied. A source of 241 Am - gamma radiation of 59.54 keV - is used. Fluorescent intensifying screens reduce the radiation dose in radiology since they produce visible light which increases the efficiency of the film. In addition, secondary X-rays arise due to the photoelectric effect, elastic (Rayleigh) and inelastic (Compton) scattering

  17. Studying atomic-resolution by X-ray fluorescence holography

    International Nuclear Information System (INIS)

    Gao Hongyi; Chen Jianwen; Xie Honglan; Zhu Huafeng; Li Ruxin; Xu Zhizhan

    2005-01-01

    In this work, the results of numerical simulations of X-ray fluorescence holograms and the reconstructed atomic images for Fe single crystal are given. The influences of the recording angles ranges and the polarization effect on the reconstruction of the atomic images are discussed. The process for removing twin images by multiple energy fluorescence holography and expanding the energy range of the incident X-rays to improve the resolution of the reconstructed images is presented

  18. X-ray fluorescent elemental analysis. Ch. 16

    International Nuclear Information System (INIS)

    Baryshev, V.; Kulipanov, G.; Skrinsky, A.

    1991-01-01

    X-ray fluorescence analysis (XFA) is used worldwide to define a quantitative content of the elements as well as to visualize the distribution of elements in different regions (element mapping). Utilization of synchrotron radiation (SR) to excite X-ray fluorescence enables the XFA method to be qualitatively improved. This chapter reviews the experimental work in especially the last decade (author). 71 refs.; 24 figs.; 3 tabs

  19. X-ray fluorescence imaging with synchrotron radiation

    International Nuclear Information System (INIS)

    Rivers, M.L.

    1987-01-01

    The micro-distribution of trace elements is of great interest in fields such as geochemistry, biology and material science. The synchrotron x-ray fluorescence microprobe provides a technique to quantitatively measure trace element compositions at individual points and to construct semiquantitative two dimensional maps of trace element compositions. This paper describes an x-ray fluorescence system used at the National Synchrotron Light Source

  20. X-ray fluorescence cross sections for K and L x rays of the elements

    International Nuclear Information System (INIS)

    Krause, M.O.; Nestor, C.W. Jr.; Sparks, C.J. Jr.; Ricci, E.

    1978-06-01

    X-ray fluorescence cross sections are calculated for the major x rays of the K series 5 less than or equal to Z less than or equal to 101, and the three L series 12 less than or equal to Z less than or equal to 101 in the energy range 1 to 200 keV. This calculation uses Scofield's theoretical partical photoionization cross sections, Krause's evaluation of fluorescence and Coster-Kronig yields, and Scofield's theoretical radiative rates. Values are presented in table and graph format, and an estimate of their accuracy is made. The following x rays are considered: Kα 1 , Kα 1 , 2 , Kβ 1 , Kβ 1 , 3 , Lα 1 , Lα 1 , 2 , Lβ 1 , Lβ 2 , 15 , Lβ 3 , Ll, Lγ 1 , Lγ 4 , and L 1 → L 2 , 3 . For use in x-ray fluorescence analysis, Kα and Lα fluorescence cross sections are presented at specific energies: TiK identical with 4.55 keV, CrK identical with 5.46 keV, CoK identical with 7.00 keV, CuK identical with 8.13 keV, MoKα identical with 17.44 keV, AgK identical with 22.5 keV, DyK identical with 47.0 keV, and 241 Am identical with 59.54 keV. Supplementary material includes fluorescence and Coster--Kronig yields, fractional radiative rates, fractional fluorescence yields, total L-shell fluorescence cross sections, fluorescence and Coster-Kronig yields in condensed matter, effective fluorescence yields, average L-shell fluorescence yield, L-subshell photoionization cross section ratios, and conversion factors from barns per atom to square centimeters per gram

  1. Polarized X-ray excitation for scatter reduction in X-ray fluorescence computed tomography.

    Science.gov (United States)

    Vernekohl, Don; Tzoumas, Stratis; Zhao, Wei; Xing, Lei

    2018-05-25

    X-ray fluorescence computer tomography (XFCT) is a new molecular imaging modality which uses X-ray excitation to stimulate the emission of fluorescent photons in high atomic number contrast agents. Scatter contamination is one of the main challenges in XFCT imaging which limits the molecular sensitivity. When polarized X-rays are used, it is possible to reduce the scatter contamination significantly by placing detectors perpendicular to the polarization direction. This study quantifies scatter contamination for polarized and unpolarized X-ray excitation and determines the advantages of scatter reduction. The amount of scatter in preclinical XFCT is quantified in Monte Carlo simulations. The fluorescent X-rays are emitted isotropically, while scattered X-rays propagate in polarization direction. The magnitude of scatter contamination is studied in XFCT simulations of a mouse phantom. In this study, the contrast agent gold is examined as an example but a scatter reduction from polarized excitation is also expected for other elements. The scatter reduction capability is examined for different polarization intensities with a monoenergetic X-ray excitation energy of 82 keV. The study evaluates two different geometrical shapes of CZT detectors which are modeled with an energy resolution of 1 keV FWHM at an X-ray energy of 80 keV. Benefits of a detector placement perpendicular to the polarization direction are shown in iterative and analytic image reconstruction including scatter correction. The contrast to noise ratio (CNR) and the normalized mean square error (NMSE) are analyzed and compared for the reconstructed images. A substantial scatter reduction for common detector sizes was achieved for 100% and 80% linear polarization while lower polarization intensities provide a decreased scatter reduction. By placing the detector perpendicular to the polarization direction, a scatter reduction by factor up to 5.5 can be achieved for common detector sizes. The image

  2. X-ray fluorescence (XRF) set-up with a low power X-ray tube

    International Nuclear Information System (INIS)

    Gupta, Sheenu; Deep, Kanan; Jain, Lalita; Ansari, M.A.; Mittal, Vijay Kumar; Mittal, Raj

    2010-01-01

    The X-ray fluorescence set-up with a 100 W X-ray tube comprises a computer controlled system developed for remote operation and monitoring of tube and an adjustable stable 3D arrangement to procure variable excitation energies with low scattered background. The system was tested at different filament currents/anode voltages. The MDL of the set-up at 0.05-1.00 mA/4-12 kV is found ∼(1-100) ppm for K and L excitations and ∼(200-700) ppm for M excitations of elements and improves with filament current and anode voltage. Moreover, L measurements for Sm and Eu at five K X-ray energies of elements(Z=29-40) and analytical determination in some synthetic samples were undertaken.

  3. The Mapping X-ray Fluorescence Spectrometer (MapX)

    Science.gov (United States)

    Sarrazin, P.; Blake, D. F.; Marchis, F.; Bristow, T.; Thompson, K.

    2017-12-01

    Many planetary surface processes leave traces of their actions as features in the size range 10s to 100s of microns. The Mapping X-ray Fluorescence Spectrometer (MapX) will provide elemental imaging at 100 micron spatial resolution, yielding elemental chemistry at a scale where many relict physical, chemical, or biological features can be imaged and interpreted in ancient rocks on planetary bodies and planetesimals. MapX is an arm-based instrument positioned on a rock or regolith with touch sensors. During an analysis, an X-ray source (tube or radioisotope) bombards the sample with X-rays or alpha-particles / gamma-rays, resulting in sample X-ray Fluorescence (XRF). X-rays emitted in the direction of an X-ray sensitive CCD imager pass through a 1:1 focusing lens (X-ray micro-pore Optic (MPO)) that projects a spatially resolved image of the X-rays onto the CCD. The CCD is operated in single photon counting mode so that the energies and positions of individual X-ray photons are recorded. In a single analysis, several thousand frames are both stored and processed in real-time. Higher level data products include single-element maps with a lateral spatial resolution of 100 microns and quantitative XRF spectra from ground- or instrument- selected Regions of Interest (ROI). XRF spectra from ROI are compared with known rock and mineral compositions to extrapolate the data to rock types and putative mineralogies. When applied to airless bodies and implemented with an appropriate radioisotope source for alpha-particle excitation, MapX will be able to analyze biogenic elements C, N, O, P, S, in addition to the cations of the rock-forming elements >Na, accessible with either X-ray or gamma-ray excitation. The MapX concept has been demonstrated with a series of lab-based prototypes and is currently under refinement and TRL maturation.

  4. Total-reflection x-ray fluorescence with a brillant undulator x-ray source

    International Nuclear Information System (INIS)

    Sakurai, K.; Eba, H.; Numako, C.; Suzuki, M.; Inoue, K.; Yagi, N.

    2000-01-01

    Total-reflection x-ray fluorescence (TXRF) is a highly sensitive technique for analyzing trace elements, because of the very low background from the sample support. Use of third-generation synchrotron x-ray source could further enhance the detection power. However, while such high sensitivity permits the detection of signals from trace elements of interest, it also means that one can observe weak parasitic x-rays as well. If the sample surface becomes even slightly contaminated, owing to air particulates near the beamline, x-ray fluorescence lines of iron, zinc, copper, nickel, chromium, and titanium can be observed even for a blank sample. Another critical problem is the low-energy-side tail of the scattering x-rays, which ultimately restricts the detection capability of the technique using a TXRF spectrometer based on a Si(Li) detector. The present paper describes our experiments with brilliant undulator x-ray beams at BL39XU and BL40XU, at the SPring-8, Harima, Japan. The emphasis is on the development of instruments to analyze a droplet of 0.1 μl containing trace elements of ppb level. Although the beamline is not a clean room, we have employed equipment for preparing a clean sample and also for avoiding contamination during transferring the sample into the spectrometer. We will report on the successful detection of the peak from 0.8 ppb selenium in a droplet (absolute amount 80 fg). We will also present the results of recent experiments obtained from a Johansson spectrometer rather than a Si(Li) detector. (author)

  5. Application of X-ray fluorescence analysis in environmental research

    International Nuclear Information System (INIS)

    Kliment, V.; Kliman, J.; Turzo, I.

    1978-01-01

    A description is presented of the X-ray fluorescence analysis principles and of its possibilities in the study of environmental pollution impact. Experiments with X-ray fluorescence analysis using 241-Am and a Ge(Li) semiconductor detector are discussed. The reproducibility of determinations in dependence on the sample preparation and the evaluation of peak surfaces of characteristic radiation is shown. The dependence of the peak surface on the elemental contents in the sample was linear. Detection limits of the investigated elements ranged in tenths of μg for 300 s measurement. (author)

  6. X-ray fluorescence microtomography analyzing prostate tissues

    International Nuclear Information System (INIS)

    Pereira, Gabriela R.; Rocha, Henrique S.; Calza, Cristiane; Lopes, Ricardo T.

    2009-01-01

    The objective of this work is to determine the elemental distribution map in reference samples and prostate tissue samples using X-Ray Fluorescence Microtomography (XRFCT) in order to verify concentrations of certain elements correlated with characteristics observed by the transmission microtomography. The experiments were performed at the X-Ray Fluorescence Facility of the Brazilian Synchrotron Light Laboratory. A quasi-monochromatic beam produced by a multilayer monochromator was used as an incident beam. The transmission CT images were reconstructed using filtered-back-projection algorithm, and the XRFCT images were reconstructed using filtered-back-projection algorithm with absorption corrections. (author)

  7. Quantitative X ray analysis system. User's manual and guide to X ray fluorescence technique

    International Nuclear Information System (INIS)

    2009-01-01

    This guide covers trimmed and re-arranged version 3.6 of the Quantitative X ray Analysis System (QXAS) software package that includes the most frequently used methods of quantitative analysis. QXAS is a comprehensive quantitative analysis package that has been developed by the IAEA through research and technical contracts. Additional development has also been carried out in the IAEA Laboratories in Seibersdorf where QXAS was extensively tested. New in this version of the manual are the descriptions of the Voigt-profile peak fitting, the backscatter fundamental parameters' and emission-transmission methods of chemical composition analysis, an expanded chapter on the X ray fluorescence physics, and completely revised and increased number of practical examples of utilization of the QXAS software package. The analytical data accompanying this manual were collected in the IAEA Seibersdorf Laboratories in the years 2006/2007

  8. MSL Chemistry and Mineralogy X-Ray Diffraction X-Ray Fluorescence (CheMin) Instrument

    Science.gov (United States)

    Zimmerman, Wayne; Blake, Dave; Harris, William; Morookian, John Michael; Randall, Dave; Reder, Leonard J.; Sarrazin, Phillipe

    2013-01-01

    This paper provides an overview of the Mars Science Laboratory (MSL) Chemistry and Mineralogy Xray Diffraction (XRD), X-ray Fluorescence (XRF) (CheMin) Instrument, an element of the landed Curiosity rover payload, which landed on Mars in August of 2012. The scientific goal of the MSL mission is to explore and quantitatively assess regions in Gale Crater as a potential habitat for life - past or present. The CheMin instrument will receive Martian rock and soil samples from the MSL Sample Acquisition/Sample Processing and Handling (SA/SPaH) system, and process it utilizing X-Ray spectroscopy methods to determine mineral composition. The Chemin instrument will analyze Martian soil and rocks to enable scientists to investigate geophysical processes occurring on Mars. The CheMin science objectives and proposed surface operations are described along with the CheMin hardware with an emphasis on the system engineering challenges associated with developing such a complex instrument.

  9. Submicron hard X-ray fluorescence imaging of synthetic elements.

    Science.gov (United States)

    Jensen, Mark P; Aryal, Baikuntha P; Gorman-Lewis, Drew; Paunesku, Tatjana; Lai, Barry; Vogt, Stefan; Woloschak, Gayle E

    2012-04-13

    Synchrotron-based X-ray fluorescence microscopy (XFM) using hard X-rays focused into sub-micron spots is a powerful technique for elemental quantification and mapping, as well as microspectroscopic measurements such as μ-XANES (X-ray absorption near edge structure). We have used XFM to image and simultaneously quantify the transuranic element plutonium at the L(3) or L(2)-edge as well as Th and lighter biologically essential elements in individual rat pheochromocytoma (PC12) cells after exposure to the long-lived plutonium isotope (242)Pu. Elemental maps demonstrate that plutonium localizes principally in the cytoplasm of the cells and avoids the cell nucleus, which is marked by the highest concentrations of phosphorus and zinc, under the conditions of our experiments. The minimum detection limit under typical acquisition conditions with an incident X-ray energy of 18 keV for an average 202 μm(2) cell is 1.4 fg Pu or 2.9×10(-20) moles Pu μm(-2), which is similar to the detection limit of K-edge XFM of transition metals at 10 keV. Copper electron microscopy grids were used to avoid interference from gold X-ray emissions, but traces of strontium present in naturally occurring calcium can still interfere with plutonium detection using its L(α) X-ray emission. Copyright © 2012 Elsevier B.V. All rights reserved.

  10. Instrument and method for X-ray diffraction, fluorescence, and crystal texture analysis without sample preparation

    Science.gov (United States)

    Gendreau, Keith (Inventor); Martins, Jose Vanderlei (Inventor); Arzoumanian, Zaven (Inventor)

    2010-01-01

    An X-ray diffraction and X-ray fluorescence instrument for analyzing samples having no sample preparation includes a X-ray source configured to output a collimated X-ray beam comprising a continuum spectrum of X-rays to a predetermined coordinate and a photon-counting X-ray imaging spectrometer disposed to receive X-rays output from an unprepared sample disposed at the predetermined coordinate upon exposure of the unprepared sample to the collimated X-ray beam. The X-ray source and the photon-counting X-ray imaging spectrometer are arranged in a reflection geometry relative to the predetermined coordinate.

  11. Energy Dispersive X-Ray Fluorescence Spectrometric Study of ...

    African Journals Online (AJOL)

    MBI

    2017-06-11

    Jun 11, 2017 ... Compositional Differences in Trace Elements in Dried Moringa oleifera ... Ti, Cu, Mo, Fe, Zn, Ni, Re, Eu and Pb using Energy Dispersive X-ray fluorescence ... Africa, Southeast Asia (Valdez-Solana et al., 2015). ... vegetable in many countries, including Nigeria .... of other elements in environmental samples.

  12. X-ray fluorescence spectrometry applied to soil analysis

    International Nuclear Information System (INIS)

    Salvador, Vera Lucia Ribeiro; Sato, Ivone Mulako; Scapin Junior, Wilson Santo; Scapin, Marcos Antonio; Imakima, Kengo

    1997-01-01

    This paper studies the X-ray fluorescence spectrometry applied to the soil analysis. A comparative study of the WD-XRFS and ED-XRFS techniques was carried out by using the following soil samples: SL-1, SOIL-7 and marine sediment SD-M-2/TM, from IAEA, and clay, JG-1a from Geological Survey of Japan (GSJ)

  13. X-ray fluorescence spectrometry - an introduction course

    International Nuclear Information System (INIS)

    Salvador, V.L.R.

    1989-01-01

    The theoretical and experimental principles of the X-ray fluorescence spectrometry, is presented the text is a synthesis of the most important literature in this area. The authors included are: E.P. Bertin, R. Jenkins, J.L. Devries, R. Muller, R. Tertian, F. Claisse e K.L. Willians. (author)

  14. A portable tube exciting X-ray fluorescence analysis system

    International Nuclear Information System (INIS)

    Yang Qiang; Lai Wanchang; Ge Liangquan

    2009-01-01

    Article introduced a portable tube exciting X-ray fluorescence analysis system which is based on arm architecture. Also, we designed Tube control circuit and finished preliminary application. The energy and the intensity of the photon can be adjusted continuously by using the tube. Experiments show that high excitation efficiency obtained by setting the appropriate parameters of the tube for the various elements. (authors)

  15. Limestone rocks analysis by X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Izquierdo M, G.; Ponce R, R.; Vazquez J, J.

    1996-01-01

    By request of a private company, employing basically X-ray fluorescence analysis (X RF), was established a fast and accurate method for the analysis of the major elements in limestone rocks. Additionally, for complementing analysis was determined by ion chromatography, the chlorides appearance and by atomic absorption of sodium. By gravimetry, was determined the losses by ignition and the alpha quartz. (Author)

  16. X-ray fluorescence spectroscopic determination of heavy metals and ...

    African Journals Online (AJOL)

    Purpose: To determine the heavy metal and trace element composition of the powdered aerial parts of Origanum sipyleum L. and its water extract. Methods: The heavy metal and trace elements content of the powdered plant material and 2 % aqueous extract were evaluated by x-ray fluorescence spectroscopy with silicon ...

  17. Research Note: Energy dispersive x-ray fluorescence analysis ...

    African Journals Online (AJOL)

    Energy Dispersive X-Ray fluorescence (EDXRF) technique for the analysis of geological, biological and environmental samples is described. The technique has been applied in the analysis of 10 (geological, biological, environmental) standard reference materials. The accuracy and precision of the technique were attested ...

  18. Measuring and interpreting X-ray fluorescence from planetary surfaces.

    Science.gov (United States)

    Owens, Alan; Beckhoff, Burkhard; Fraser, George; Kolbe, Michael; Krumrey, Michael; Mantero, Alfonso; Mantler, Michael; Peacock, Anthony; Pia, Maria-Grazia; Pullan, Derek; Schneider, Uwe G; Ulm, Gerhard

    2008-11-15

    As part of a comprehensive study of X-ray emission from planetary surfaces and in particular the planet Mercury, we have measured fluorescent radiation from a number of planetary analog rock samples using monochromatized synchrotron radiation provided by the BESSY II electron storage ring. The experiments were carried out using a purpose built X-ray fluorescence (XRF) spectrometer chamber developed by the Physikalisch-Technische Bundesanstalt, Germany's national metrology institute. The XRF instrumentation is absolutely calibrated and allows for reference-free quantitation of rock sample composition, taking into account secondary photon- and electron-induced enhancement effects. The fluorescence data, in turn, have been used to validate a planetary fluorescence simulation tool based on the GEANT4 transport code. This simulation can be used as a mission analysis tool to predict the time-dependent orbital XRF spectral distributions from planetary surfaces throughout the mapping phase.

  19. X-ray imaging with monochromatic synchrotron radiation. Fluorescent and phase-contrast method

    Energy Technology Data Exchange (ETDEWEB)

    Takeda, Tohoru; Itai, Yuji [Tsukuba Univ., Ibaraki (Japan). Inst. of Clinical Medicine

    2002-05-01

    To obtain the high sensitive x-ray images of biomedical object, new x-ray imaging techniques using fluorescent x-ray and phase-contrast x-ray are being developed in Japan. Fluorescent x-ray CT can detect very small amounts of specific elements in the order of ppm at one pixel, whereas phase-contrast x-ray imaging with interferometer can detect minute differences of biological object. Here, our recent experimental results are presented. (author)

  20. Investigation of Pink Tourmalines by X-ray Fluorescent Technique

    International Nuclear Information System (INIS)

    Sangariyavanich, A.; Na Songkhla, S.; Pimjum, S.

    1998-01-01

    X-ray fluorescent technique has been employed in the study of trace elements in six samples of gamma irradiated pink tourmalines, namely, red-pink (rubellite), light-pink, orange-pink, brownish orange-pink, purple red and purple orange-pink. The analysis of their characteristic X-ray indicated the existence of manganese in all samples. Trace amounts of iron, zinc, lead, bismuth or gallium were also investigated in certain samples. Since these elements were not present in red-pink tourmaline, therefore, we believed that manganese is the major cause of pink color in tourmaline while other elements produce various types of pink color

  1. Effect of Electric Voltage and Current of X-ray Chamber on the Element inthe Zirconium Alloy Analysis X-ray by X-ray Fluorescence

    International Nuclear Information System (INIS)

    Yusuf-Nampira; Narko-Wibowo, L; Rosika-Krisnawati; Nudia-Barenzani

    2000-01-01

    The using of x-ray fluorescence in the chemical analysis depend heavilyon the parameters of x-ray chamber, for examples : electric voltage andelectric current. That parameter give effect in the result of determine ofSn, Cr, Fe and Ni in the zirconium alloy. 20 kV electric voltages are used onthe Mo x-ray chamber shall product x-ray of zirconium in the sample materialcan give effect in the stability of the analysis result (deviation more than5%). The result of analysis of elements in the zirconium alloy shall givedeviation less than 5% when using of electric voltage of the x-ray chamberless than 19 kV. The sensitivity of analysis can be reached by step upelectric current of x-ray chamber. (author)

  2. Extra-oral dental radiography for disaster victims using a flat panel X-ray detector and a hand-held X-ray generator.

    Science.gov (United States)

    Ohtani, M; Oshima, T; Mimasaka, S

    2017-12-01

    Forensic odontologists commonly incise the skin for post-mortem dental examinations when it is difficult to open the victim's mouth. However, it is prohibited by law to incise dead bodies without permission in Japan. Therefore, we attempted using extra-oral dental radiography, using a digital X-ray equipment with rechargeable batteries, to overcome this restriction. A phantom was placed in the prone position on a table, and three plain dental radiographs were used per case: "lateral oblique radiographs" for left and right posterior teeth and a "contact radiograph" for anterior teeth were taken using a flat panel X-ray detector and a hand-held X-ray generator. The resolving power of the images was measured by a resolution test chart, and the scattered X-ray dose was measured using an ionization chamber-type survey meter. The resolving power of the flat panel X-ray detector was 3.0 lp/mm, which was less than that of intra-oral dental methods, but the three extra-oral plain dental radiographs provided the overall dental information from outside of the mouth, and this approach was less time-consuming. In addition, the higher dose of scattered X-rays was laterally distributed, but the dose per case was much less than that of intra-oral dental radiographs. Extra-oral plain dental radiography can be used for disaster victim identification by dental methods even when it is difficult to open the mouth. Portable and rechargeable devices, such as a flat panel X-ray detector and a hand-held X-ray generator, are convenient to bring and use anywhere, even at a disaster scene lacking electricity and water.

  3. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens

    International Nuclear Information System (INIS)

    Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K.

    2004-01-01

    A polycapillary X-ray lens is an effective optics to obtain a μm-size X-ray beam for micro-X-ray fluorescence spectrometry (μ-XRF). We developed a μ-XRF instrument using a polycapillary X-ray lens, which also enabled us to perform Grazing Exit μ-XRF (GE-μ-XRF). The evaluated diameter of the primary X-ray beam was 48 μm at the focal distance of the X-ray lens. Use of this instrument enabled two-dimensional mapping of the elemental distributions during growth of the plant 'Quinoa'. The results of the mapping revealed elemental transition during growth. In addition, a small region of thin film was analyzed by GE-μ-XRF. We expect that GE-μ-XRF will become an effective method of estimating the film thickness of a small region

  4. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens

    Energy Technology Data Exchange (ETDEWEB)

    Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K. E-mail: tsuji@a-chem.eng.osaka-cu.ac.jp

    2004-08-31

    A polycapillary X-ray lens is an effective optics to obtain a {mu}m-size X-ray beam for micro-X-ray fluorescence spectrometry ({mu}-XRF). We developed a {mu}-XRF instrument using a polycapillary X-ray lens, which also enabled us to perform Grazing Exit {mu}-XRF (GE-{mu}-XRF). The evaluated diameter of the primary X-ray beam was 48 {mu}m at the focal distance of the X-ray lens. Use of this instrument enabled two-dimensional mapping of the elemental distributions during growth of the plant 'Quinoa'. The results of the mapping revealed elemental transition during growth. In addition, a small region of thin film was analyzed by GE-{mu}-XRF. We expect that GE-{mu}-XRF will become an effective method of estimating the film thickness of a small region.

  5. Analyses of archaeological pottery samples using X-ray fluorescence technique for provenance study

    International Nuclear Information System (INIS)

    Tamilarasu, S.; Swain, K.K.; Singhal, R.K; Reddy, A.V.R.; Acharya, R.; Velraj, G.

    2015-01-01

    Archaeological artifacts reveal information on past human activities, artifact preparation technology, art and possible trade. Ceramics are the most stable and abundant material in archaeological context. Pottery is the most abundant tracers in all archaeological excavations. Compared to major elements, elements present at trace concentrations levels are source specific and they maintain same concentration levels in source clay as well as finished products e.g., fired clay potteries. As it is difficult to find out exact source or origin, provenance study is carried out first to establish whether objects under study are from the same or different sources/origin. Various analytical techniques like instrumental neutron activation analysis (INAA), Ion beam analysis (IBA) and X-ray fluorescence (XRF) have been used for obtaining elemental concentrations in archaeological potteries. Portable X-ray fluorescence (pXRF) spectrometry provides a non-destructive means for elemental characterization of a wide range of archaeological materials. Ten archaeological pottery samples were collected from Kottapuram, Kerala under the supervision of archaeological survey of India. Portable X-ray fluorescence (pXRF) spectrometry using a handheld Olympus Innov-X Delta XRF device, ACD BARC, has been used for chemical characterization of the pottery samples. The instrument is equipped with the Delta Rhodium (Rh) anode X-Ray tube and uses a Silicon Drift Detector (resolution <200 eV at 5.95 keV Mn Kα X-ray). NIST 2781 SRM was analyzed for quality control purpose. Ten elements namely Fe, Ti, Mn, Co, Cu, Zn, Pb, Zr, Mo and Se were chosen for cluster analysis and their concentration values were utilized for multivariate statistical analysis using WinSTAT 9.0

  6. Monitoring body iron burden using X-ray fluorescence (XRF)

    International Nuclear Information System (INIS)

    Farquharson, M.J.; Bagshaw, A.P.

    2001-01-01

    X-ray fluorescence, using Cu K alpha and K beta radiation, has been used to measure the Fe content of skin of two groups of rats, one Fe overloaded and one control group. These skin Fe levels were compared to the liver and heart Fe levels measured using colorimetry. Correlation coefficients of 0.86 and 0.88 respectively were found indicating that skin Fe levels may be a potential marker for body iron burden.

  7. Method and device for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Jagoutz, E.; Palme, C.

    1978-01-01

    In the x-ray fluorescence analyzer the useful signal can be completely separated from the spurious signals, and especially the pulse can be determined. For this purpose the output of the radiation detector is connected with a multichannel pulse height discriminator. The measured signal determined in the pulse heigth discriminator may be indicated by a visual display or processed by a computer (coincidence circuits). (DG) [de

  8. Portable x-ray fluorescence spectrometer for Works of art

    International Nuclear Information System (INIS)

    Mendoza, A.; Griesser, A.

    2001-01-01

    X-ray fluorescence is an analytical technique of prier importance in archaeometry, for restoration and art history investigation; it is because of non-destructive and multi-elemental character of the analysis simplicity and high speed of operation, ability to produce immediate analytical results for the objects, which can neither be sampled nor removed to the laboratory Recent advances in X-ray tubes, X-ray detectors and electronic provided an opportunity to produce portable high resolution XRF spectrometers characterized by a good reliability and analytical performance; in this paper a prototype portable XRF spectrometer based on a small size, low power X-ray tube and a thermometrically cooled Si-Pin detector is described. The spectrometer provides a possibility for direct and secondary target excitation geometry use of proper secondary target and filter and size adjustment of the primary photon bean by using a set of different beam collimators; the portable XRF spectrometer was successfully applied to study art objects in the Art History Museum in Vienna, including such objects as old master paintings bronze and brass alloys of antique as well as Renaissance objects and silver/copper coins produced at different locations. Quantitative and Quantitative analysis were amedee depending of the curator questions and discussed from the point of view of art History. The importance of the results for restoration and authentification of the art objects is also emphasized

  9. Relative probabilities of the uranium isotopes for thorium x-ray emission and fluorescence of uranium x-rays

    International Nuclear Information System (INIS)

    Parker, J.L.

    1991-01-01

    Both thorium x-rays from decaying uranium isotopes and self-fluoresced uranium x-rays are prominent in high-resolution gamma-ray spectra of uranium-bearing materials. Useful application of the information carried by those x-rays has been curtailed because the probabilities of the uranium isotopes for thorium x-ray emission and for uranium x-ray fluorescence have not been known. By analyzing enrichment-meter geometry spectra from uranium oxide standards whose enrichments ranged from 0.7% to 91%, relative values, primarily, have been obtained for the probabilities of both processes. Thorium x-ray emission is very heavily dominated by 235 U. In all ordinarily occurring uranium isotopic distributions, thorium x-rays may be used as a valid 235 U signature. The probability for a thorium K α1 x-ray to be emitted in the decay of a 235 U atom is 0.048 ±0.002. In infinitely thick uranium oxide materials, the relative ratios of effectiveness for self-fluorescence, on a per unit mass basis, are approximately 234 U : 235 U : 236 U : 238 U = 1.13 : 1.00 : 0.52 : 0.028. on a per decay basis, the approximate ratios are 0.00039 : 1.00 : 0.017 : 0.18. These results imply that, contrary to what has often been stated, gamma rays are far more important than alpha particles in the self-fluorescence of uranium. Because of the importance of gamma-ray self-fluorescence, the uranium x-ray yield will be somewhat influenced by the size, shape, and composition of the materials. 4 refs., 1 fig

  10. Quantitative x-ray fluorescent analysis using fundamental parameters

    International Nuclear Information System (INIS)

    Sparks, C.J. Jr.

    1976-01-01

    A monochromatic source of x-rays for sample excitation permits the use of pure elemental standards and relatively simple calculations to convert the measured fluorescent intensities to an absolute basis of weight per unit weight of sample. Only the mass absorption coefficients of the sample for the exciting and the fluorescent radiation need be determined. Besides the direct measurement of these absorption coefficients in the sample, other techniques are considered which require fewer sample manipulations and measurements. These fundamental parameters methods permit quantitative analysis without recourse to the time-consuming process of preparing nearly identical standards

  11. Trends in environmental science using microscopic X-ray fluorescence

    International Nuclear Information System (INIS)

    Fittschen, Ursula Elisabeth Adriane; Falkenberg, Gerald

    2011-01-01

    Microscopic X-ray fluorescence (micro-XRF) is a versatile tool in environmental analysis. We review work done in this field from 2008 to 2010 and highlight new aspects. Overall, there is a strong trend to combine fluorescence data with other data like diffraction or absorption spectroscopy. Also, the use of laboratory based instrumentation has become wide spread as more commercial instruments are available. At laboratories and synchrotron sites the trend towards higher spatial resolution is still persistent hitting sub micrometer values in case of synchrotron set ups.

  12. Trends in environmental science using microscopic X-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Fittschen, Ursula Elisabeth Adriane, E-mail: ursula.fittschen@chemie.uni-hamburg.de [Department of Chemistry, University of Hamburg, Martin-Luther-King-Platz 6, 20146 Hamburg (Germany); Falkenberg, Gerald [Deutsches Elektronen-Synchrotron, Notkestr. 85, 22603 Hamburg (Germany)

    2011-08-15

    Microscopic X-ray fluorescence (micro-XRF) is a versatile tool in environmental analysis. We review work done in this field from 2008 to 2010 and highlight new aspects. Overall, there is a strong trend to combine fluorescence data with other data like diffraction or absorption spectroscopy. Also, the use of laboratory based instrumentation has become wide spread as more commercial instruments are available. At laboratories and synchrotron sites the trend towards higher spatial resolution is still persistent hitting sub micrometer values in case of synchrotron set ups.

  13. X-ray scattering in X-ray fluorescence spectra with X-ray tube excitation - Modelling, experiment, and Monte-Carlo simulation

    International Nuclear Information System (INIS)

    Hodoroaba, V.-D.; Radtke, M.; Vincze, L.; Rackwitz, V.; Reuter, D.

    2010-01-01

    X-ray scattering may contribute significantly to the spectral background of X-ray fluorescence (XRF) spectra. Based on metrological measurements carried out with a scanning electron microscope (SEM) having attached a well characterised X-ray source (polychromatic X-ray tube) and a calibrated energy dispersive X-ray spectrometer (EDS) the accuracy of a physical model for X-ray scattering is systematically evaluated for representative samples. The knowledge of the X-ray spectrometer efficiency, but also of the spectrometer response functions makes it possible to define a physical spectral background of XRF spectra. Background subtraction relying on purely mathematical procedures is state-of-the-art. The results produced by the analytical model are at least as reliable as those obtained by Monte-Carlo simulations, even without considering the very challenging contribution of multiple scattering. Special attention has been paid to Compton broadening. Relevant applications of the implementation of the analytical model presented in this paper are the prediction of the limits of detection for particular cases or the determination of the transmission of X-ray polycapillary lenses.

  14. Absorption correction factor in X-ray fluorescent quantitative analysis

    International Nuclear Information System (INIS)

    Pimjun, S.

    1994-01-01

    An experiment on absorption correction factor in X-ray fluorescent quantitative analysis were carried out. Standard samples were prepared from the mixture of Fe 2 O 3 and tapioca flour at various concentration of Fe 2 O 3 ranging from 5% to 25%. Unknown samples were kaolin containing 3.5% to-50% of Fe 2 O 3 Kaolin samples were diluted with tapioca flour in order to reduce the absorption of FeK α and make them easy to prepare. Pressed samples with 0.150 /cm 2 and 2.76 cm in diameter, were used in the experiment. Absorption correction factor is related to total mass absorption coefficient (χ) which varied with sample composition. In known sample, χ can be calculated by conveniently the formula. However in unknown sample, χ can be determined by Emission-Transmission method. It was found that the relationship between corrected FeK α intensity and contents of Fe 2 O 3 in these samples was linear. This result indicate that this correction factor can be used to adjust the accuracy of X-ray intensity. Therefore, this correction factor is essential in quantitative analysis of elements comprising in any sample by X-ray fluorescent technique

  15. Gaseous detectors for energy dispersive X-ray fluorescence analysis

    Science.gov (United States)

    Veloso, J. F. C. A.; Silva, A. L. M.

    2018-01-01

    The energy resolution capability of gaseous detectors is being used in the last years to perform studies on the detection of characteristic X-ray lines emitted by elements when excited by external radiation sources. One of the most successful techniques is the Energy Dispersive X-ray Fluorescence (EDXRF) analysis. Recent developments in the new generation of micropatterned gaseous detectors (MPGDs), triggered the possibility not only of recording the photon energy, but also of providing position information, extending their application to EDXRF imaging. The relevant features and strategies to be applied in gaseous detectors in order to better fit the requirements for EDXRF imaging will be reviewed and discussed, and some application examples will be presented.

  16. Quantitative analysis by computer controlled X-ray fluorescence spectrometer

    International Nuclear Information System (INIS)

    Balasubramanian, T.V.; Angelo, P.C.

    1981-01-01

    X-ray fluorescence spectroscopy has become a widely accepted method in the metallurgical field for analysis of both minor and major elements. As encountered in many other analytical techniques, the problem of matrix effect generally known as the interelemental effects is to be dealt with effectively in order to make the analysis accurate. There are several methods by which the effects of matrix on the analyte are minimised or corrected for and the mathematical correction is one among them. In this method the characteristic secondary X-ray intensities are measured from standard samples and correction coefficients. If any, for interelemental effects are evaluated by mathematical calculations. This paper describes attempts to evaluate the correction coefficients for interelemental effects by multiple linear regression programmes using a computer for the quantitative analysis of stainless steel and a nickel base cast alloy. The quantitative results obtained using this method for a standard stainless steel sample are compared with the given certified values. (author)

  17. Fluorescence versus X-ray cholangiography during laparoscopic cholecystectomy

    DEFF Research Database (Denmark)

    Lehrskov, Lars Lang; Larsen, Søren S; Kristensen, Billy B

    2016-01-01

    INTRODUCTION: Intraoperative fluorescent cholangiography is a novel non-invasive imaging technique to visualise the extrahepatic biliary tract during laparoscopic cholecystectomy. It has been proven feasible, fast and cost effective. Never-theless, there is only sparse data on the capacity...... of fluorescent cholangiography to visualise the biliary anatomy. METHODS: Based on a non-inferiority design, patients with complicated gallstone disease are randomised to either -intraoperative conventional X-ray cholangiography (reference group, n = 60) or intraoperative fluorescent cholangiography (n = 60......). The primary outcome is visualisation of the junction between the cystic duct, the common hepatic duct and the common bile duct. CONCLUSION: The present study may show that fluorescent cholangiography is as valid for visualisation of important structures of the extrahepatic biliary tract as conventional X...

  18. Analysis of fresco paintings by X-ray fluorescence method

    International Nuclear Information System (INIS)

    Cechak, T.; Gerndt, J.; Musilek, L.; Kopecka, I.

    2000-01-01

    In this work we present the application of X-ray fluorescence analysis (XRFA) to examine fresco paintings from the Karlstejn castle. The X-ray fluorescence apparatus built and operated in the Laboratory of Quantitative Methods in Research of Ancient Monuments was used for the purpose of fresco paintings measurements. The X-ray sources (radionuclides) generate the characteristic X-ray photons from the sample. The Si(Li) detector measures numbers and energies of photons emitted from the specimen. The energy and number of photons detected can be converted into kind and amount of measured atoms. These results give data for qualitative and quantitative analysis of samples. XRFA is relatively simple and non-destructive method. Capability of in-situ measurement is one of big advantages of this method. The radionuclide sources of exciting radiation (e.g. 55 Fe enables the excitation of elements with Z up to 23, 238 Pu is used in interval of Z from 20 to 39 etc.) were used. An Si(Li) semiconductor detector with a 5 l Dewar vessel and portable spectroscopy system enable the in situ measurement. Narrow collimation of the exciting beam makes it possible to select the measured area of fresco painting. The valuable fresco paintings from the Karlstejn castle were investigated in this way. The measurements were carried out in collaboration with the Analytical Laboratory of the State Institute for the Preservation of Historic Monuments. A suitable analysis of paintings makes it possible to detect the kind of colours and evaluate changes in the surface colour of paintings and suggest useful and timely procedures for their conservation and restoration. (author)

  19. A new method for x-ray fluorescence analysis of contaminated material. Final Report

    International Nuclear Information System (INIS)

    Grodzins, Lee; Niland, John

    2002-01-01

    Niton has successfully completed the objectives of the Phase II program to build a hand-held, x-ray fluorescent analyzer optimized for DOE decontamination and decommissioning activities in the field. A two-pound x-ray fluorescence analyzer was developed that contains 3 radioactive sources, emitting 3 widely spaced monochromatic x-rays, to give the lowest detection limits for the full range of toxic elements, from chromium to plutonium. A rapid, fundamental- parameters algorithm was developed that yields quantitative results in less than 1 second. High-resolution silicon drift detectors and silicon PIN diodes give excellent efficiency and speed. These results from Phase II have been introduced into the XL 300, 700 and 800 commercial products series. More than 800 of these instruments, yielding revenues of more than $20 million dollars, have been sold since the first 3-source instrument was introduced in 1998. A direct consequence of the Phase II funding has been the growth of Niton from 20 people to its present size of 60

  20. A new method for x-ray fluorescence analysis of contaminated material. Final Report

    Energy Technology Data Exchange (ETDEWEB)

    Grodzins, Lee; Niland, John

    2002-05-23

    Niton has successfully completed the objectives of the Phase II program to build a hand-held, x-ray fluorescent analyzer optimized for DOE decontamination and decommissioning activities in the field. A two-pound x-ray fluorescence analyzer was developed that contains 3 radioactive sources, emitting 3 widely spaced monochromatic x-rays, to give the lowest detection limits for the full range of toxic elements, from chromium to plutonium. A rapid, fundamental- parameters algorithm was developed that yields quantitative results in less than 1 second. High-resolution silicon drift detectors and silicon PIN diodes give excellent efficiency and speed. These results from Phase II have been introduced into the XL 300, 700 and 800 commercial products series. More than 800 of these instruments, yielding revenues of more than $20 million dollars, have been sold since the first 3-source instrument was introduced in 1998. A direct consequence of the Phase II funding has been the growth of Niton from 20 people to its present size of 60.

  1. X-ray Peltier cooled detectors for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Loupilov, A.; Sokolov, A.; Gostilo, V.

    2001-01-01

    The recent results on development of X-ray Si(Li), Si-planar and CdTe p-i-n detectors cooled by Peltier coolers for fabrication of laboratory and portable XRF analysers for different applications are discussed. Low detection limits of XRF analysers are provided by increasing of detectors sensitive surface; improvement of their spectrometrical characteristics; decreasing of front-end-electronics noise level; Peltier coolers and vacuum chambers cooling modes optimization. Solution of all mentioned tasks allowed to develop Peltier cooled detectors with the following performances: (1.) Si(Li) detectors: S=20 mm 2 , thickness=3.5 mm, 175 eV (5.9 keV), 430 eV (59.6 keV); S=100 mm 2 ; thickness=4.5 mm, 270 eV (5.9 keV), 485 eV (59.6 keV). (2.) Si-planar detector: S=10 mm 2 , thickness=0.4 mm, 230 eV (5.9 keV), 460 eV (59.6 keV). (3.) CdTe p-i-n detectors: S=16 mm 2 , thickness=0.5 mm, 350 eV (5.9 keV), 585 eV (59.6 keV). S=16 mm 2 , thickness=1.2 mm, 310 eV (5.9 keV), 600 eV (59.6 keV). Advantages and disadvantages of all types of detectors for X-ray fluorescence analysis are compared. Spectra are presented. Application of different XRF analysers based on developed detectors in medicine, environmental science, industry, cryminalistics and history of art are demonstrated

  2. X-ray Peltier cooled detectors for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Loupilov, A.; Sokolov, A.; Gostilo, V.

    2000-01-01

    The recent results on development of X-ray Si(Li), Si-planar and CdTe p-i- n detectors cooled by Peltier coolers for fabrication of laboratory and portable XRF analysers for different applications are discussed. Low detection limits of XRF analysers are provided by increasing of detectors sensitive surface; improvement of their spectrometrical characteristics; decreasing of front-end-electronics noise level; Peltier coolers and vacuum chambers cooling modes optimization. Solution of all mentioned tasks allowed to develop Peltier cooled detectors with the following performances: (1) Si(Li) detectors: S = 20 mm 2 , thickness = 3.5 mm, 175 eV (5.9 keV), 430 eV (59.6 keV); S = 100 mm 2 ; thickness = 4.5 mm, 270 eV (5.9 keV), 485 eV (59,6 keV). (2) Si-planar detector: S = 10 mm 2 , thickness = 0.4 mm, 230 eV (5.9 keV), 460 eV (59.6 keV). (3) CdTe p-i-n detectors: S = 16 mm 2 , thickness 0.5 mm, 350 eV (5.9 keV), 585 eV (59.6 keV). S = 16 mm 2 , thickness = 1.2 mm, 310 eV (5.9 keV), 600 eV (59.6 keV). Advantages and disadvantages of all types of detectors for X-ray fluorescence analysis are compared. Spectra are presented. Application of different XRF analysers based on developed detectors in medicine, environmental science, industry, criminalistics and history of art are demonstrated. (author)

  3. Simulation and application of micro X-ray fluorescence based on an ellipsoidal capillary

    Science.gov (United States)

    Yang, Jing; Li, Yude; Wang, Xingyi; Zhang, Xiaoyun; Lin, Xiaoyan

    2017-06-01

    A micro X-ray fluorescence setup was presented, based on an ellipsoidal capillary and a traditional laboratorial X-ray source. Using Ray-tracing principle, we have simulated the transmission path of X-ray beam in the ellipsoidal capillary and designed the optimal parameters of the ellipsoidal capillary for the micro X-ray fluorescence setup. We demonstrate that ellipsoidal capillary is well suited as condenser for the micro X-ray fluorescence based on traditional laboratorial X-ray source. Furthermore, we obtain the 2D mapping image of the leaf blade sample by using the ellipsoidal capillary we designed.

  4. X-ray fluorescence in IAEA Member States: Spain

    International Nuclear Information System (INIS)

    Roldan, C.; Ferrero, J.L.

    2004-01-01

    Full text: Instrumental facilities of the ICMUV include: a Total-reflection X-Ray Fluorescence (TXRF), laboratory and portable Energy Dispersive X-Ray Fluorescence (EDXRF) spectrometers. These equipments are employed in the field of the art and archaeometry. Current projects are: EDXRF analysis of blue pigments used in Valencian ceramics. EDXRF analyses of cobalt-blue pigments were made on 73 pieces of Valencian ceramics from the beginning of the 14th century up to 20th century. These ceramic samples have the pigment decoration applied together with a tin opacified lead glaze cover on the clay body. The comparison between EDXRF spectra from coloured and non-coloured areas provides information about the pigment composition. The following elements: Mn, Fe, Co, Ni, Cu, Zn and As are identified as characteristics of the blue pigments. Different association of these elements as well as correlation with the chronology of the samples were found. These results can be used for identifying the different types of cobalt ores employed in the manufacture of the blue pigments to study their provenance. Non-destructive analysis of paper supports used in prints: In paper based works of art it is not possible to separate the support from the work of the author. Then, the maximum knowledge of the support in this kind of works is desirable. In this work, Energy Dispersive X-Ray Fluorescence (EDXRF) was used to determine the elemental composition of a set of European and Oriental papers from the 20th century and an Arabian paper from the 14th century. These papers were manufactured with different production techniques and used as support for writing, drawing and printing. Normalised fluorescence yields of the elements to the weight of the paper show that there are some correlations between its elemental composition and the type of paper, provenance and use. Therefore, the Energy Dispersive X-Ray Fluorescence (EDXRF) technique could be used for a better characterization and

  5. High resolution projection X-ray microscope equipped with fluorescent X-ray analyzer and its applications

    International Nuclear Information System (INIS)

    Minami, K; Saito, Y; Kai, H; Shirota, K; Yada, K

    2009-01-01

    We have newly developed an open type fine-focus X-ray tube 'TX-510' to realize a spatial resolution of 50nm and to radiate low energy characteristic X-rays for giving high absorption contrast to images of microscopic organisms. The 'TX-510' employs a ZrO/W(100) Schottky emitter and an 'In-Lens Field Emission Gun'. The key points of the improvements are (1) reduced spherical aberration coefficient of magnetic objective lens, (2) easy and accurate focusing, (3) newly designed astigmatism compensator, (4) segmented thin film target for interchanging the target materials by electron beam shift and (5) fluorescent X-ray analysis system.

  6. X-ray fluorescence holography and multiple-energy x-ray holography: A critical comparison of atomic images

    International Nuclear Information System (INIS)

    Len, P.M.; Gog, T.; Fadley, C.S.; Materlik, G.

    1997-01-01

    We compare x-ray fluorescence holography (XFH) and multiple-energy x-ray holography (MEXH), two techniques that have recently been used to obtain experimental three-dimensional atomic images. For single-energy holograms, these methods are equivalent by virtue of the optical reciprocity theorem. However, XFH can only record holographic information at the characteristic fluorescence energies of the emitting species, while MEXH can record holographic information at any energy above the fluorescent edge of the emitter, thus enabling the suppression of real-twin overlaps and other aberrations and artifacts in atomic images. copyright 1997 The American Physical Society

  7. Rare earth aerosol analysis by x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Citron, I.M.; Mausner, L.F.

    1982-01-01

    An analytical method for the determination of four lanthanides in air filter samples is described. The method involves simultaneous quantitative determinations of La, Ce, Pr, and Nd at the microgram level by x-ray fluorescence spectrometry without chemical separation of these rare earths and without serious interferences from the dust matrices on the filters. The method has been used successfully to analyze some air filter samples collected at a rare earth processing refinery in Illinois. A description of the development of the method is given as well as the results obtained by using this method on the air filter samples. The reproducibility of the results was generally +-5%

  8. Improved fluorescent X-ray image intensifying screen

    International Nuclear Information System (INIS)

    Landeghem, W.K. van; Suys, A.R.

    1981-01-01

    An X-ray image intensifying screen is described, which includes at least one fluorescent layer comprising phosphor particles dispersed in a binder and on top of such layer a protective layer containing a crosslinked polymer mass obtained by an acid-catalyzed reaction of a polymer or mixture of polymers containing reactive hydrogen atoms and a cross-linking agent, the cross-linking agent being an organic compound containing a plurality of etherified N-methylol groups. Examples are given of appropriate polymers and cross-linking agents. (author)

  9. X-ray fluorescence analysis of welding fume particles

    International Nuclear Information System (INIS)

    Carsey, T.P.

    1982-01-01

    A commercial standard filter set and two laboratory-made standard filter sets are compared via the analysis of generated welding fume samples by X-ray fluorescence. The latter standards are made by (1) hydrophobic-edge membrane filters spiked with prepared metal ion solutions, and (2) filters through which a dispersion of metal oxide powder in isopropanol has been drawn. The results are presented in table form. Precision (Pre) is the relative standard deviation of the six samples. Four main conclusions are enumerated

  10. Blood selenium content determination by X-ray fluorescence

    International Nuclear Information System (INIS)

    Mainardi, R.T.

    1987-01-01

    The presence of some elements in small amounts (traces) in the human body is of foremost importance for the prevention and treatment of several diseases. It has been recently shown that traces of selenium in blood are closely related to the occurrence of miotonic distrophy, a muscular disease that is affecting a significant percentage of the population. This work describes a simple procedure to determine selenium in human blood serum by energy dispersive X-ray fluorescence analysis. Final quantification is achieved through the addition of titanium as an internal standard. (Author) [es

  11. Studies of some alloys using x-ray fluorescence

    International Nuclear Information System (INIS)

    Elmahi, Elamin Musaid

    1997-01-01

    In this project an attempt has been made for the study of alloys commonly used using x-ray fluorescence ( XRF ) technique. The alloys selected for the study included gold jewellery, steels, brasses and coins. The XRF method proved to be simple, fast, non-destructive and reliable as compared to chemical methods. The results showed that most of the gold jewellery used in this country have carat value of 18 and 21. Also most coins used in different countries are alloys of Cu and Ni. A simple spark method was used for the determination of C in steels, since C is not possible to analyze by XRF. ( Author )

  12. Sensitometric curve of radiographic films by X-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Gonçalves, E.A.S., E-mail: elicardo.goncalves@ifrj.edu.br [Instituto Federal do Rio de Janeiro (IFRJ), RJ (Brazil); Oliveira, D.F.; Anjos, M.J. dos; Oliveira, L.F. de [Universidade do Estado do Rio de Janeiro (UERJ), RJ (Brazil); Lopes, R.T. [Coordenacao de Pos-Graduacao e Pesquisa de Engenharia (COPPE/UFRJ), Rio de Janeiro, RJ (Brazil)

    2017-07-01

    Radiographic film exposure is traditionally measured by the transmittance of a beam of light through the film. There are many mathematical and computational models to characterize the curve behavior and its properties, but almost none of them considers the limitations caused by the equipment used. As long as exposure in film increases, light intensity measured after the film decreases in a way that from a certain exposure, light could not be distinguished from any kind of noise. This work aims to propose x-ray fluorescence as a solution for better measure high exposed films and show how it could be modeled mathematically. (author)

  13. Quantification procedures in micro X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Kanngiesser, Birgit

    2003-01-01

    For the quantification in micro X-ray fluorescence analysis standardfree quantification procedures have become especially important. An introduction to the basic concepts of these quantification procedures is given, followed by a short survey of the procedures which are available now and what kind of experimental situations and analytical problems are addressed. The last point is extended by the description of an own development for the fundamental parameter method, which renders the inclusion of nonparallel beam geometries possible. Finally, open problems for the quantification procedures are discussed

  14. X-Ray Diffraction and Fluorescence Instrument for Mineralogical Analysis at the Lunar Surface, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — We propose to develop LUNA, a compact and lightweight X-Ray Diffraction (XRD) / X-Ray Fluorescence (XRF) instrument for mineralogical analysis of regolith, rock...

  15. X-Ray Diffraction and Fluorescence Instrument for Mineralogical Analysis at the Lunar Surface, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — We propose to develop a compact and lightweight X-Ray Diffraction (XRD) / X-Ray Fluorescence (XRF) instrument for analysis of mineralogical composition of regolith,...

  16. A low power x-ray tube for use in energy dispersive x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Kataria, S.K.; Govil, Rekha; Lal, M.

    1980-01-01

    A low power X-ray tube with thin molybdenum transmission target for use in energy dispersive X-ray fluorescence (ENDXRF) element analysis has been indigenously built, along with its power supply. The X-ray tube has been in operation since August 1979, and it has been operated upto maximum target voltage of 35 KV and tube current upto 200 μA which is more than sufficient for trace element analysis. This X-ray tube has been used alongwith the indigenously built Si(Li) detector X-ray spectrometer with an energy resolution of 200 eV at 5.9 Kev MnKsub(α) X-ray peak for ENDXRF analysis. A simple procedure of calibration has been developed for thin samples based on the cellulose diluted, thin multielement standard pellets. Analytical sensitivities of the order of a few p.p.m. have been obtained with the experimental setup for elements with 20 < = Z < = 38 and 60 < = Z < = 90. A number of X-ray spectra for samples of environmental, biological, agricultural, industrial and metallurgical interest are presented to demonstrate the salient features of the experimental sep up. (auth.)

  17. Total reflection x-ray fluorescence - an approach to nanoanalysis

    International Nuclear Information System (INIS)

    Klockenkaemper, R.

    2000-01-01

    X-ray fluorescence analysis (XRFA) is a powerful tool used for industrial production, geological prospecting and for environmental control. However, the method suffers from a lack of sensitivity so that analyses are restricted to microanalytical investigations. That means: the sample amount needed for analysis is above some 10 micrograms, concentrations to be determined have to be on the μg/ml level, and thin layers to be characterized must be of micrometer thickness. In contrast to conventional XRFA, total-reflection X-ray fluorescence (TXRF) is extremely sensitive and even allows nano-analytical investigations. Three different ways can be taken: (i) use of minute sample amounts of only 10 nano-grams, (ii) determination of extreme traces below ng/ml and (iii) surface analysis and depth profiling of shallow layers with nano-meter thickness. In this lecture, the basic physical phenomena of total reflection and standing waves are outlined. The experimental equipment for TXRF is sketched out and commercially available instruments of different manufacturers are compared. Furthermore, examples are given for the three kinds of nano-analytical applications: ultra-micro, analysis, ultra trace analysis and mono- and thin-layer analysis. (author)

  18. Preparation of tissue samples for X-ray fluorescence microscopy

    International Nuclear Information System (INIS)

    Chwiej, Joanna; Szczerbowska-Boruchowska, Magdalena; Lankosz, Marek; Wojcik, Slawomir; Falkenberg, Gerald; Stegowski, Zdzislaw; Setkowicz, Zuzanna

    2005-01-01

    As is well-known, trace elements, especially metals, play an important role in the pathogenesis of many disorders. The topographic and quantitative elemental analysis of pathologically changed tissues may shed some new light on processes leading to the degeneration of cells in the case of selected diseases. An ideal and powerful tool for such purpose is the Synchrotron Microbeam X-ray Fluorescence technique. It enables the carrying out of investigations of the elemental composition of tissues even at the single cell level. The tissue samples for histopathological investigations are routinely fixed and embedded in paraffin. The authors try to verify the usefulness of such prepared tissue sections for elemental analysis with the use of X-ray fluorescence microscopy. Studies were performed on rat brain samples. Changes in elemental composition caused by fixation in formalin or paraformaldehyde and embedding in paraffin were examined. Measurements were carried out at the bending magnet beamline L of the Hamburger Synchrotronstrahlungslabor HASYLAB in Hamburg. The decrease in mass per unit area of K, Br and the increase in P, S, Fe, Cu and Zn in the tissue were observed as a result of the fixation. For the samples embedded in paraffin, a lower level of most elements was observed. Additionally, for these samples, changes in the composition of some elements were not uniform for different analyzed areas of rat brain

  19. Preparation of tissue samples for X-ray fluorescence microscopy

    Energy Technology Data Exchange (ETDEWEB)

    Chwiej, Joanna [Faculty of Physics and Applied Computer Science, AGH-University of Science and Technology, Al. Mickiewicza 30, 30-059 Cracow (Poland)]. E-mail: jchwiej@novell.ftj.agh.edu.pl; Szczerbowska-Boruchowska, Magdalena [Faculty of Physics and Applied Computer Science, AGH-University of Science and Technology, Al. Mickiewicza 30, 30-059 Cracow (Poland); Lankosz, Marek [Faculty of Physics and Applied Computer Science, AGH-University of Science and Technology, Al. Mickiewicza 30, 30-059 Cracow (Poland); Wojcik, Slawomir [Faculty of Physics and Applied Computer Science, AGH-University of Science and Technology, Al. Mickiewicza 30, 30-059 Cracow (Poland); Falkenberg, Gerald [Hamburger Synchrotronstrahlungslabor at Deutsches Elektronen-Synchrotron, Notkestr. 85, Hamburg (Germany); Stegowski, Zdzislaw [Faculty of Physics and Applied Computer Science, AGH-University of Science and Technology, Al. Mickiewicza 30, 30-059 Cracow (Poland); Setkowicz, Zuzanna [Department of Neuroanatomy, Institute of Zoology, Jagiellonian University, Ingardena 6, 30-060 Cracow (Poland)

    2005-12-15

    As is well-known, trace elements, especially metals, play an important role in the pathogenesis of many disorders. The topographic and quantitative elemental analysis of pathologically changed tissues may shed some new light on processes leading to the degeneration of cells in the case of selected diseases. An ideal and powerful tool for such purpose is the Synchrotron Microbeam X-ray Fluorescence technique. It enables the carrying out of investigations of the elemental composition of tissues even at the single cell level. The tissue samples for histopathological investigations are routinely fixed and embedded in paraffin. The authors try to verify the usefulness of such prepared tissue sections for elemental analysis with the use of X-ray fluorescence microscopy. Studies were performed on rat brain samples. Changes in elemental composition caused by fixation in formalin or paraformaldehyde and embedding in paraffin were examined. Measurements were carried out at the bending magnet beamline L of the Hamburger Synchrotronstrahlungslabor HASYLAB in Hamburg. The decrease in mass per unit area of K, Br and the increase in P, S, Fe, Cu and Zn in the tissue were observed as a result of the fixation. For the samples embedded in paraffin, a lower level of most elements was observed. Additionally, for these samples, changes in the composition of some elements were not uniform for different analyzed areas of rat brain.

  20. Determination of technetium by total reflection x-ray fluorescence

    International Nuclear Information System (INIS)

    Bermudez, J.I.; Greaves, E.D.; Nemeth, P.

    2000-01-01

    We describe a technique using total reflection x-ray fluorescence (TXRF) for determination of Technetium produced by elution of chromatography generators with physiological saline solutions. The analysis with the 18.41 keV K α line of Technetium was accomplished with monochromatized K α radiation from a silver anode x-ray tube operated at 45 keV and 20 mA. This radiation at 22.104 keV is efficiently coupled to the 21.054 keV absorption edge of Tc. It is also of advantage in the direct analysis of organic and saline properties of the Tc-bearing samples. Quantification was accomplished by internal standard addition of Ga and using an interpolated value of the sensitivity for Tc between Molybdenum and Rhenium. Data processing was carried out with the QXAS-AXIL software package. System sensitivity was found adequate for direct Tc determination of eluted saline solutions. The interest and advantages of the use of the technique as an auxiliary in the synthesis and characterization of Tc-labeled radiopharmaceuticals used for diagnosis in nuclear medicine are discussed. Detection limits in the matrices analyzed are reported. (author)

  1. Automating X-ray Fluorescence Analysis for Rapid Astrobiology Surveys.

    Science.gov (United States)

    Thompson, David R; Flannery, David T; Lanka, Ravi; Allwood, Abigail C; Bue, Brian D; Clark, Benton C; Elam, W Timothy; Estlin, Tara A; Hodyss, Robert P; Hurowitz, Joel A; Liu, Yang; Wade, Lawrence A

    2015-11-01

    A new generation of planetary rover instruments, such as PIXL (Planetary Instrument for X-ray Lithochemistry) and SHERLOC (Scanning Habitable Environments with Raman Luminescence for Organics and Chemicals) selected for the Mars 2020 mission rover payload, aim to map mineralogical and elemental composition in situ at microscopic scales. These instruments will produce large spectral cubes with thousands of channels acquired over thousands of spatial locations, a large potential science yield limited mainly by the time required to acquire a measurement after placement. A secondary bottleneck also faces mission planners after downlink; analysts must interpret the complex data products quickly to inform tactical planning for the next command cycle. This study demonstrates operational approaches to overcome these bottlenecks by specialized early-stage science data processing. Onboard, simple real-time systems can perform a basic compositional assessment, recognizing specific features of interest and optimizing sensor integration time to characterize anomalies. On the ground, statistically motivated visualization can make raw uncalibrated data products more interpretable for tactical decision making. Techniques such as manifold dimensionality reduction can help operators comprehend large databases at a glance, identifying trends and anomalies in data. These onboard and ground-side analyses can complement a quantitative interpretation. We evaluate system performance for the case study of PIXL, an X-ray fluorescence spectrometer. Experiments on three representative samples demonstrate improved methods for onboard and ground-side automation and illustrate new astrobiological science capabilities unavailable in previous planetary instruments. Dimensionality reduction-Planetary science-Visualization.

  2. X-ray fluorescence spectrometry and related techniques an introduction

    CERN Document Server

    Margui, Eva

    2013-01-01

    X-ray fluorescence spectrometry (XRF) is a well-established analytical technique for qualitative and quantitative elemental analysis of a wide variety of routine quality control and research samples. Among its many desirable features, it delivers true multi-element character analysis, acceptable speed and economy, easy of automation, and the capacity to analyze solid samples. This remarkable contribution to this field provides a comprehensive and up-to-date account of basic principles, recent developments, instrumentation, sample preparation procedures, and applications of XRF analysis. If you are a professional in materials science, analytic chemistry, or physics, you will benefit from not only the review of basics, but also the newly developed technologies with XRF. Those recent technological advances, including the design of low-power micro- focus tubes and novel X-ray optics and detectors, have made it possible to extend XRF to the analysis of low-Z elements and to obtain 2D or 3D information on a microme...

  3. X-ray fluorescence and the study of microcirculation

    International Nuclear Information System (INIS)

    Muthuvelu, P.; Hugtenburg, R.P.; Bradley, D.A.; Winlove, C.P.

    2004-01-01

    The feasibility of using K-shell X-ray fluorescence (XRF) technique for study of subchondral bone microcirculation in ex vivo samples is examined. Studies have been carried out at the Daresbury Laboratory Synchrotron Radiation Source (SRS) ultra-dilute extended X-ray absorption fine structure beamline. Initial investigations were made on fine-bore capillaries with diameters of either 500 or 200 μm, attenuated by up to 2 mm of Perspex and containing dilute iodine-based contrast media. This allowed comparison to be made with the capabilities of angiographic imaging systems, also allowing definition of suitable XRF set-up parameters for subsequent microcirculation studies. Measurements were obtained in 30 s run times, for concentrations of iodine (K ab 33.164 keV) down to 2 /(ml saline). Intensities were linear up to 3.7 mg/ml, self-absorption becoming significant for concentrations beyond this. To determine detection limits, preliminary studies of subchondral bone microcirculation were made on bone sections which were known to be poorly infused with silver-coated (K ab 25.517 keV) 30 μm diameter microspheres. For a 2 mm slice of bone, the presence of small numbers of silver-coated microspheres were detected in the first 2 mm layer from the surface, at a level equivalent to ∼1 ppm of silver solution

  4. Human thyroid specimen imaging by fluorescent x-ray computed tomography with synchrotron radiation

    Science.gov (United States)

    Takeda, Tohoru; Yu, Quanwen; Yashiro, Toru; Yuasa, Tetsuya; Hasegawa, Yasuo; Itai, Yuji; Akatsuka, Takao

    1999-09-01

    Fluorescent x-ray computed tomography (FXCT) is being developed to detect non-radioactive contrast materials in living specimens. The FXCT system consists of a silicon (111) channel cut monochromator, an x-ray slit and a collimator for fluorescent x ray detection, a scanning table for the target organ and an x-ray detector for fluorescent x-ray and transmission x-ray. To reduce Compton scattering overlapped on the fluorescent K(alpha) line, incident monochromatic x-ray was set at 37 keV. The FXCT clearly imaged a human thyroid gland and iodine content was estimated quantitatively. In a case of hyperthyroidism, the two-dimensional distribution of iodine content was not uniform, and thyroid cancer had a small amount of iodine. FXCT can be used to detect iodine within thyroid gland quantitatively and to delineate its distribution.

  5. Simulation and application of micro X-ray fluorescence based on an ellipsoidal capillary

    Energy Technology Data Exchange (ETDEWEB)

    Yang, Jing; Li, Yude; Wang, Xingyi; Zhang, Xiaoyun; Lin, Xiaoyan, E-mail: yangjing_928@126.com

    2017-06-15

    Highlights: • A micro X-ray fluorescence setup based on an ellipsoidal capillary was presented. • The optimal parameters of ellipsoidal capillary were designed. • The 2D mapping image of biological sample was obtained. - Abstract: A micro X-ray fluorescence setup was presented, based on an ellipsoidal capillary and a traditional laboratorial X-ray source. Using Ray-tracing principle, we have simulated the transmission path of X-ray beam in the ellipsoidal capillary and designed the optimal parameters of the ellipsoidal capillary for the micro X-ray fluorescence setup. We demonstrate that ellipsoidal capillary is well suited as condenser for the micro X-ray fluorescence based on traditional laboratorial X-ray source. Furthermore, we obtain the 2D mapping image of the leaf blade sample by using the ellipsoidal capillary we designed.

  6. X-ray fluorescent scanning of the thyroid

    International Nuclear Information System (INIS)

    Jonckheer, M.H.; Deconinck, F.

    1983-01-01

    The main emphasis of the technical chapters of this monograph lies on the aspects which are of direct importance to thyroid scanning: the general principles of X-ray fluorescence, the choice and characteristics of appropriate sources and detectors, a stationary system, quantification problems, and the pitfalls in the interpretation of the intrathyroidal iodine imaging and quantification. The clinical part of the monograph consists of chapters on the role of stable iodine and the thyroid function, on endemic non-toxic goiter, on hyperthyroidism as a result of iodine overload, on feasibility of dynamic studies, on stable iodine stores in thyroiditis, and on a general review of the clinical usefulness of XRF in thyroid disease. (Auth.)

  7. Analysis of solar blocker through portable X-ray fluorescence

    International Nuclear Information System (INIS)

    Ferreira, Diego de Dio; Melquiades, Fabio Luiz; Appoloni, Carlos Roberto; Lopes, Fabio; Lonni, Audrey Stinghen G.; Oliveira, Frederico Minardi de; Duarte, Jose C.

    2009-01-01

    This paper estimates the concentration of TiO 2 by Energy Dispersion X-ray fluorescence (EDXRF) viewing t obtain the FPS due to the physical barrier in the composition of solar blockers, and identifies possible present metals in the samples. A portable EDXRF equipment was used and 27 commercial of different brands and solar protection factors were analysed. Also, three formulations (A, B and C) were prepared and measured estimated in FPS-30 using 5% or TiO 2 . The quantification was performed through calibration curves with 1% to 30% standards of TiO 2 . As result, it was possible to determine the contribution to physical protection in the FPS, associated to the Ti concentration present in some solar blocker samples available in the market. Also, it was possible to detect the presence of various metals in solar protectors, such as Fe, Zn, Br and Sr, and identify chemical elements which were not mentioned and their formulation as well

  8. X-ray fluorescence in research on Czech cultural monuments

    International Nuclear Information System (INIS)

    Cechak, T.; Gerndt, J.; Kopecka, I.; Musilek, L.

    2004-01-01

    Radionuclide X-ray fluorescence analysis (RXRFA), as a non-destructive method, relatively simple and allowing measurements to be carried out in situ, is an excellent tool in research on various objects of art. A range of artefacts have been investigated by our laboratory, in part for the purposes of history of art and in part as a basis for restoration works - medieval frescoes in some Czech castles and churches, metal sculptures and objects of applied art, paints and inks of old manuscripts, paintings. Some of these are among the most valuable monuments in the Czech cultural heritage. The contribution of the results of the tests to the investigation of their 'life story' and, in some cases, to their conservation, is not negligible. Later additions and repairs can be recognised, and materials and technologies that are close to their historic counterparts can be used in restoration work

  9. Uranium concentrate analysis by X-ray fluorescence spectroscopy

    International Nuclear Information System (INIS)

    Diaz-Guerra, J.P.; Bayon, A.; Roca, R.

    1978-01-01

    The determination of As, Ca, Fe, Mo, P, S, Si. Th, V and U in uranium concentrates by X-ray fluorescence spectroscopy has been studied. As and U are determined in nitric solutions and for the rest of elements analysis is performed by a bead fusion technique using Li 2 B 4 O 7 and Li 2 CO 3 as fluxes. Although the uranium matrix minimizes the absorption and enhancement effects, because of the content variations of this element it is advisable to operate at a constant level of U 3 O 8 . Despite the high matrix absorption and the large dilution of the samples, sensitivity and speed are found to be satisfactory as the result of the use of a high sensitivity automatic spectrometer. The spectral interferences of Mo on S and P, and of Pb on As have been particularly considered. (author) [es

  10. Environmental studies in Khartoum area using x-ray fluorescence

    International Nuclear Information System (INIS)

    Abdel Elmagid, Suliman Alamin

    1996-06-01

    In the present work an attempt has been made for the analysis of some soil, plant, sediments and fish samples of relevance to environmental pollution in Khartoum area. These samples have been collected from different places in residential areas, so as to cover industrial areas, agricultural and residential areas, as well as Tuti Island as control area. Special attention has been dedicated to the analysis of lead concentrations resulting from automobile-emissions in soils and to other toxic metals such as Cr in some industries. The samples were analysed by x-ray fluorescence (XRF) technique. The results obtained using XRF measurements and computer software called QXAS for data analysis. The concentrations of lead and some heavy metals such as Cr in soils from certain locations were alarming and may create pollution problems in the near future. The results obtained from different countries. The results are generally lower than the international limits. (Author)

  11. Handbook of practical X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Beckhoff, B.; Wedell, R.; Wolff, H.

    2006-01-01

    X-ray fluorescence analysis (XRF) is a reliable multi-elemental and nondestructive analytical method widely used in research and industrial applications. This practical handbook provides self-contained modules featuring XRF instrumentation, quantification methods, and most of the current applications. The broad spectrum of topics is due to the efforts of a large number of authors from a variety of different types of institutions such as universities, research institutes, and companies. The book gives a survey of the theoretical fundamentals, analytical instrumentation, software for data processing, various excitation regimes including gracing incidents and microfocus measurements, quantitative analysis, applications in routine and micro analysis, mineralogy, biology, medicine, criminal investigations, archeology, metallurgy, abrasion, microelectronics, environmental air and water analysis. It gives the basic knowledge on this technique, information on analytical equipment and guides the reader to the various applications. This practical handbook is intended as a resource for graduate students, research scientists, and industrial users. (orig.)

  12. Handbook of practical X-ray fluorescence analysis

    Energy Technology Data Exchange (ETDEWEB)

    Beckhoff, B. [Physikalisch-Technische Bundesanstalt, Berlin (Germany). X-ray Spectrometry; Kanngiesser, B. [Technische Univ. Berlin (Germany). Inst. fuer Atomare Physik und Fachdidaktik; Langhoff, N. [IfG-Institute for Scientific Instruments GmbH, Berlin (Germany); Wedell, R.; Wolff, H. (eds.) [Institut fuer Angewandte Photonik e.V., Berlin (Germany)

    2006-07-01

    X-ray fluorescence analysis (XRF) is a reliable multi-elemental and nondestructive analytical method widely used in research and industrial applications. This practical handbook provides self-contained modules featuring XRF instrumentation, quantification methods, and most of the current applications. The broad spectrum of topics is due to the efforts of a large number of authors from a variety of different types of institutions such as universities, research institutes, and companies. The book gives a survey of the theoretical fundamentals, analytical instrumentation, software for data processing, various excitation regimes including gracing incidents and microfocus measurements, quantitative analysis, applications in routine and micro analysis, mineralogy, biology, medicine, criminal investigations, archeology, metallurgy, abrasion, microelectronics, environmental air and water analysis. It gives the basic knowledge on this technique, information on analytical equipment and guides the reader to the various applications. This practical handbook is intended as a resource for graduate students, research scientists, and industrial users. (orig.)

  13. Compositional analysis of steels by x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Kutty, K.V.G.; Rajagopalan, S.; Ananthakrishnan, S.K.; Asuvatharaman, R.

    1988-01-01

    The x-ray fluorescence spectrometric method of determining Ni, Cr, Mn, Mo, T i, Si and Co in austenitic, ferritic, titanium-stabilised and low alloy steels is described. Samples in the bulk solid form (25-50 mm diameter and upto 50 mm height) or very small samples having a flat surface at least 5 mm in diameter are analysed as received after a brief surface treatment. The complex matrix effects affecting the determination of the various elements are discussed. The direct calibration method as well as the mathematical method of empirical coefficients of overcoming the matrix effects is described. The precision and accuracy of the analyses are discussed. (author). 10 refs

  14. Hyper-filter-fluorescer spectrometer for fusion x-ray diagnostics

    International Nuclear Information System (INIS)

    Wang, C.L.

    1981-01-01

    The filter-fluorescer spectrometer (FFS) is a powerful tool for measuring x-ray spectrum from high fluence x-ray sources. However, this technique is limited to energies less than 120 keV, because there are no practical absorption edges available above this energy. In this paper, we present a new method of utilizing the filter-fluorescer system for x-ray spectral measurement above 120 keV. The new apparatus is called hyper-filter-fluorescer spectrometer

  15. X-ray fluorescence beamline at LNLS: components and some associated techniques

    International Nuclear Information System (INIS)

    Perez, CArlos A.; Radtke, Martin; Perez, Carlos; Tolentino, Helio; Vicentin, Flavio; Sanchez, Hector Jorge; Perez, Roberto D.

    1997-01-01

    Full text. In this work a general description of the Total Reflection X-Ray Fluorescence (TXRF) and the X-Ray Fluorescence Microprobe (XRFM) is presented. Components, equipment and experimental stations for the x-ray fluorescence beamline are described, regarding to the techniques mentioned above. Results from the simulations of a pair bended mirrors in a Kirkpatrick-Baez configuration, are shown. The simulations were performed with Shadow program. (author)

  16. Direct comparison of soft x-ray images of organelles with optical fluorescence images

    International Nuclear Information System (INIS)

    Ishino, Masahiko; Kado, Masataka; Kishimoto, Maki; Nishikino, Masaharu; Ohba, Toshiyuki; Kaihori, Takeshi; Kawachi, Tetsuya; Tamotsu, Satoshi; Yasuda, Keiko; Mikata, Yuji; Shinohara, Kunio

    2011-01-01

    Soft x-ray microscopes operating in the water window region are capable of imaging living hydrated cells. Up to now, we have been able to take some soft x-ray images of living cells by the use of a contact x-ray microscope system with laser produced plasma soft x-ray source. Since the soft x-ray images are different from the optical images obtained with an ordinary microscope, it is very important to identify what is seen in the x-ray images. Hence, we have demonstrated the direct comparison between the images of organelles obtained with a fluorescence microscope and those with a soft x-ray microscope. Comparing the soft x-ray images to the fluorescence images, the fine structures of the organelles could be identified and observed. (author)

  17. Application of synchrotron radiation to x-ray fluorescence analysis of trace elements

    International Nuclear Information System (INIS)

    Gordon, B.M.; Jones, K.W.; Hanson, A.L.

    1986-08-01

    The development of synchrotron radiation x-ray sources has provided the means to greatly extend the capabilities of x-ray fluorescence analysis for determinations of trace element concentrations. A brief description of synchrotron radiation properties provides a background for a discussion of the improved detection limits compared to existing x-ray fluorescence techniques. Calculated detection limits for x-ray microprobes with micrometer spatial resolutions are described and compared with experimental results beginning to appear from a number of laboratories. The current activities and future plans for a dedicated x-ray microprobe beam line at the National Synchrotron Light Source (NSLS) of Brookhaven National Laboratory are presented

  18. Case Studies on Facility Characterization with X-Ray Fluorescence Spectrometry

    International Nuclear Information System (INIS)

    Kirk, K.T.; Brooksbank, R.D.; Meszaros, J.M.; Towery, W.E.

    2008-01-01

    A hand-held x-ray fluorescence (XRF) analyzer is being used to characterize facilities in support of demolition activities at the East Tennessee Technology Park (ETTP) in Oak Ridge, Tennessee. Approximately 500 facilities at the U.S. Department of Energy site are being demolished under the ETTP Decontamination and Decommissioning (D and D) project. Facility characterization is being conducted to provide data for waste profiling and identify hazards to demolition workers. XRF spectrometry is a non-destructive analytical technique used to identify and quantify the elemental composition of a substance based on the intensity of its characteristic X-ray emission wavelength or energy. The Innov-X Systems R Model XT-245S XRF analyzer used at ETTP is equipped with a silver anode x-ray tube and a Si PIN diode detector. X-rays are generated by electrical current, eliminating the need for radioactive isotopes. Electronic components can be powered by either a lithium-ion battery or an A/C adapter, and the instrument is controlled by an iPAQ R pocket personal computer. The unit has two primary operating modes. Alloy analysis mode measures percent levels of elements in metals such as a pipes, valves, equipment, or construction materials. Soil mode provides parts-per-million (ppm) quantities in bulk solids like concrete dust, residue, paint chips, or soil. The hand-held unit can analyze material in place, or it can analyze samples in a test stand by remote operation. This paper present some case studies demonstrating a variety of XRF applications for facility characterization: Metal Materials Characterization, Lead Paint Identification, Hot Spot Delineation, Bulk Solids Testing. XRF has been the analytical technique of choice for identifying metal alloy components and has also been useful in analyzing bulk materials. Limitations of XRF testing include the inability to directly analyze elements with low atomic weights. Light elements such as beryllium and aluminum do not emit

  19. Development of portable X-ray diffractometer equipped with X-ray fluorescence spectrometer and its application to archaeology

    International Nuclear Information System (INIS)

    Yamashita, Daisuke; Ishizaki, Atsushi; Uda, Masayuki

    2009-01-01

    A portable X-ray diffractometer equipped with an X-ray fluorescence spectrometer was improved so as to obtain a diffraction pattern and a fluorescence spectrum in air from one and the same small area of a specimen. The reason why the portable XRD with an XRF spectrometer was specially designed for archaeology may be understood from the following facts: (1) some objects exhibited in museums are not allowed to be transferred from the open air to a vacuum, even if their volumes are small; (2) some objects are very difficult to move from their original sites; (3) some parts of exhibits are extremely fragile and cannot be examined in a vacuum; and (4) information on the chemical composition and structure from the same area of an object offers a better understanding of the constitutive materials of the object. Some examples of the use of a portable X-ray diffractometer equipped with an X-ray fluorescence spectrometer in the field are also introduced. Experimental results of Sho-kannon, Snew's mask and Tutankhamun's golden mask are shown here. (author)

  20. Background approximation in automatic qualitative X-ray-fluorescent analysis

    International Nuclear Information System (INIS)

    Jordanov, J.; Tsanov, T.; Stefanov, R.; Jordanov, N.; Paunov, M.

    1982-01-01

    An empirical method of finding the dependence of the background intensity (Isub(bg) on the wavelength is proposed, based on the approximation of the experimentally found values for the background in the course of an automatic qualitative X-ray fluorescent analysis with pre-set curve. It is assumed that the dependence I(lambda) will be well approximated by a curve of the type Isub(bg)=(lambda-lambda sub(o)sup(fsub(1)(lambda))exp[fsub(2)(lambda)] where fsub(1) (lambda) and f 2 (lambda) are linear functions with respect to the sought parameters. This assumption was checked out on a ''pure'' starch background, in which it is not known beforehand which points belong to the background. It was assumed that the dependence I(lambda) can be found from all minima in the spectrum. Three types of minima has been distinguished: 1. the lowest point between two well-solved X-ray lines; 2. a minimum obtained as a result of statistical fluctuations of the measured signal; 3. the lowest point between two overlapped lines. The minima strongly deviating from the background are removed from the obtained set. The sum-total of the remaining minima serves as a base for the approximation of the dependence I(lambda). The unknown parameters are determined by means of the LSM. The approximated curve obtained by this method is closer to the real background than the background determined by the method described by Kigaki Denki, as the effect of all recorded minima is taken into account. As an example the PbTe spectrum recorded with crystal LiF 220 is shown graphically. The curve well describes the background of the spectrum even in the regions in which there are no minima belonging to the background. (authors)

  1. Fast x-ray fluorescence microtomography of hydrated biological samples.

    Directory of Open Access Journals (Sweden)

    Enzo Lombi

    Full Text Available Metals and metalloids play a key role in plant and other biological systems as some of them are essential to living organisms and all can be toxic at high concentrations. It is therefore important to understand how they are accumulated, complexed and transported within plants. In situ imaging of metal distribution at physiological relevant concentrations in highly hydrated biological systems is technically challenging. In the case of roots, this is mainly due to the possibility of artifacts arising during sample preparation such as cross sectioning. Synchrotron x-ray fluorescence microtomography has been used to obtain virtual cross sections of elemental distributions. However, traditionally this technique requires long data acquisition times. This has prohibited its application to highly hydrated biological samples which suffer both radiation damage and dehydration during extended analysis. However, recent advances in fast detectors coupled with powerful data acquisition approaches and suitable sample preparation methods can circumvent this problem. We demonstrate the heightened potential of this technique by imaging the distribution of nickel and zinc in hydrated plant roots. Although 3D tomography was still impeded by radiation damage, we successfully collected 2D tomograms of hydrated plant roots exposed to environmentally relevant metal concentrations for short periods of time. To our knowledge, this is the first published example of the possibilities offered by a new generation of fast fluorescence detectors to investigate metal and metalloid distribution in radiation-sensitive, biological samples.

  2. Layer thickness measurement using the X-ray fluorescence principle

    International Nuclear Information System (INIS)

    Mengelkamp, B.

    1980-01-01

    Curium 244 having a gamma energy of about 15.5 keV is used as excitation emitter for contactless and continuous measuring of the thickness of metallic layers on iron strip. Soft gamma radiation is absorbed in matter according to the photo effect, so that X-ray fluorescence radiation is generated in the matter, which depends on the element and is radiated to all sides. For instance, it amounts for iron 6.4 keV and is measured with a specific ionisation chamber for this energy range. With increasing atomic number of the elements, the energy of fluorescence radiation increases and hence also the emission signal of the detector. The prerequisite for a usable measuring effect is an element distance of at least two and the thickness of the layer to be measured being in an optimum range. A signal dependent on the thickness of the layer is produced either by absorption of iron radiation (absorption method - aluminium and tin) or by build-up radiation of the material of the layer (emission method - zinc and lead). (orig./GSCH) [de

  3. Sensitivity of in vivo X-ray fluorescence determination of skeletal lead stores

    International Nuclear Information System (INIS)

    Sokas, R.K.; Besarab, A.; McDiarmid, M.A.; Shapiro, I.M.; Bloch, P.

    1990-01-01

    Eighteen patients with known past occupational lead exposure underwent parenteral diagnostic chelation with ethylenediaminetetraacetic acid and x-ray fluorescent determination of in vivo skeletal lead stores at the distal styloid process of the ulna and at the temporal base bone using a cobalt 57 source and measuring lead Ka x-rays. X-ray fluorescent lead measurements in both locations correlated with results of diagnostic chelation. Using a post-chelation urinary excretion of greater than 600 micrograms lead/24 h as the definition of high-lead stores, sensitivity of x-ray fluorescence at the wrist and temple was 56% and 39%, respectively

  4. Results of improvement of simultaneous and sequential x-ray fluorescence equipment for quantitative routine analysis

    International Nuclear Information System (INIS)

    Zsamboky, Jozsef

    1985-01-01

    Two main types of x-ray fluorescence analyzers measuring sequentially and simultaneously, respectively, the intensities at given wave lengths are described. The main parts of an up to date x-ray fluorescence analyzer are surveyed in detail. The advantages and disadvantages of both methods are discussed. Some results on calibration and optimization are given. (D.Gy.)

  5. Sulfur content measurement in coal by X-ray fluorescence method

    International Nuclear Information System (INIS)

    Cechak, T.; Thinova, L.

    2001-01-01

    X-ray fluorescence, using backscattering, was employed in the determination of sulfur content and ash content measurement in coal. The results of the methods are given to illustrate the differences between the chemical analysis and X-ray fluorescence method.

  6. Dispersive X-ray fluorescence applications in energy in environmental problems diagnostic

    International Nuclear Information System (INIS)

    Odino, R.; Souto, B.; Roca, S.; Campomar, W.

    1994-01-01

    X-ray fluorescence energy was used to detect the grade of contamination due to a Portland cement factory. X-ray fluorescence was used to determine the incidence of a Portland cement plant in the quality of air in its surroundings. Many contaminants (Cu, Pb, Ni, Br) do not come from the Portland cement industry but other industries in the zone

  7. Renal function measured by X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Boijsen, M.; Jacobsson, L.; Tylen, U.; Sahlgrenska Sjukhuset, Goeteborg

    1987-01-01

    In twenty-six patients injected with metrizoate during urography, plasma was analyzed for iodine concentration using X-ray fluorescence analysis, and total plasma clearance of contrast medium was calculated. Total plasma clearance of 51 Cr-EDTA was also determined, but not simultaneously, in order to find out if the urographic procedure would influence the kidneys to such an extent that the contrast medium clearance value would differ much from the 51 Cr-EDTA clearance value. The errors in the method were assessed and the total error of the contrast medium clearance determination was calculated. When comparing 51 Cr-EDTA and metrizoate clearance a correlation of 0.94 and a mean ratio of 1.046 (SD 0.138) was found. The assessable errors cannot fully explain the standard deviation, which might indicate a transient change in kidney function related to elements of the urographic procedure such as laxation with possible dehydration and/or the contrast medium dose. (orig.)

  8. Fluorescence x-ray analyzer for plating-bath solution

    International Nuclear Information System (INIS)

    Komatsu, Shigemi; Hato, Yoshio; Tono, Miki; Ishijima, Hiroshi

    1984-01-01

    This analyzer was developed for the analysis of plating solution composition and the measurement of plating thickness at the same time in the noble and base metal plating applied to electronic components. The analyzer operates on the principle of fluorescence X-ray measurement which features the capability of high accuracy, non-destructive and multi-element simultaneous analysis. In this paper, the principle of measurement, the configuration of a model SFA 875 analyzer and also the main specifications and measurement examples are described. As the measurement examples, it is described in detail that the model SFA 875 achieved the expansion of the range of application and the improvement of accuracy and the accuracy of simply repeated measurement by combining a digital filter with the linear method of least squares. The digital filter method is that for eliminating noise in data, smoothing spectra and subtracting background. The linear method of least squares is the method for separating the spectra with overlapping peaks into individual peaks. The combination of them allows the analyzer to perform various analyses even by using the spectra obtained with proportional counters. The analysis of the elements whose atomic numbers are close to each other is also possible. The accuracy of measurement of plating thickness is about 0.2 to 2.0 μm, and that of solution composition is about 0.4 to 0.7 g/l. (Wakatsuki, Y.)

  9. Radioisotope sources for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Leonowich, J.; Pandian, S.; Preiss, I.L.

    1977-01-01

    Problems involved in developing radioisotope sources and the characteristics of potentially useful radioisotopes for X-ray fluorescence analysis are presented. These include the following. The isotope must be evaluated for the physical and chemical forms available, purity, half-life, specific activity, toxicity, and cost. The radiation hazards of the source must be considered. The type and amount of radiation output of the source must be evaluated. The source construction must be planned. The source should also present an advance over those currently available in order to justify its development. Some of the isotopes, which are not in use but look very promising, are indicated, and their data are tabulated. A more or less ''perfect'' source within a given range of interest would exhibit the following characteristics. (1) Decay by an isometric transition with little or no internal conversion, (2) Have an intense gamma transition near the absorption edge of the element(s) of interest with no high energy gammas, (3) Have a sufficiently long half-life (in the order of years) for both economic and calibration reasons, (4) Have a sufficiently large cross-section for production in a reasonable amount of time. If there are competing reactions the interfering isotopes should be reasonably short-lived, or if not, be apt to be separated from the isotope chemically with a minimum of difficulty. (T.G.)

  10. Thin film characterization by total reflection x-ray fluorescence

    International Nuclear Information System (INIS)

    Danel, Adrien; Nolot, Emmanuel; Veillerot, Marc; Olivier, Segolene; Decorps, Tifenn; Calvo-Munoz, Maria-Luisa; Hartmann, Jean-Michel; Lhostis, Sandrine; Kohno, Hiroshi; Yamagami, Motoyuki; Geoffroy, Charles

    2008-01-01

    Sensitive and accurate characterization of films thinner than a few nm used in nanoelectronics represents a challenge for many conventional production metrology tools. With capabilities in the 10 10 at/cm 2 , methods usually dedicated to contamination analysis appear promising, especially Total-reflection X-Ray Fluorescence (TXRF). This study shows that under usual configuration for contamination analysis, with incident angle smaller than the critical angle of the substrate, TXRF signal saturation occurs very rapidly for dense films (below 0.5 nm for HfO 2 films on Si wafers using a 9.67 keV excitation at 0.5 deg.). Increasing the incident angle, the range of linear results can be extended, but on the other hand, the TXRF sensitivity is degraded because of a strong increase of the measurement dead time. On HfO 2 films grown on Si wafers, an incident angle of 0.32 deg. corresponding to a dead time of 95% was used to achieve linear analysis up to 2 nm. Composition analysis by TXRF, and especially the detection of minor elements into thin films, requires the use of a specific incident angle to optimize sensitivity. Although quantitative analyses might require specific calibration, this work shows on Co-based films that the ratio between minor elements (W, P, Mo) and Co taking into account their relative sensitivity factors is a good direct reading of the composition

  11. A Brazilian tree collection analyzed by X ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, H; Fernandes, E A.N.; Ferraz, E S.B. [Centro de Energia Nuclear na Agricultura, Universidade de Sao Paulo, Laboratorio de Radioisotopos, Piracicaba (Brazil); Haselberger, N; Markowicz, A [International Atomic Energy Agency, Seibersdorf (Austria)

    2000-06-01

    The analysis of the inorganic components of wood is of great interest for several reasons, including the acquisition of basic data creating a data base of values for individual species. Knowing the wide variability in matrix composition (lignin, oil, resin, silica) and densities (0.39-1.09 g cm{sup -3}), 40 species of trees were analysed by X ray fluorescence (XRF) to determine the concentrations of Br, Ca, Cu, K, Mn, Pb, Rb, Sr, and A. This technique is widely used because of its accuracy and simplicity of sample preparation, normally complex for this type of biological material. This multi-elemental analysis has proven suitable for wood, a material in which a wide range for each element was encountered in the different species studied: 0.3-5.2 for Br, 126-9074 for Ca, 2.2-11 for Cu, 108-5873 for K, 3.1-134 for Mn, 0.5-4.7 for Pb, 0.3-20 for Rb, 1.2-120 for Sr, and 1.1-20 for Zn (values given in {mu}g g{sup -1}). (author)

  12. A Brazilian tree collection analyzed by X ray fluorescence

    International Nuclear Information System (INIS)

    Oliveira, H.; Fernandes, E.A.N.; Ferraz, E.S.B.; Haselberger, N.; Markowicz, A.

    2000-01-01

    The analysis of the inorganic components of wood is of great interest for several reasons, including the acquisition of basic data creating a data base of values for individual species. Knowing the wide variability in matrix composition (lignin, oil, resin, silica) and densities (0.39-1.09 g cm -3 ), 40 species of trees were analysed by X ray fluorescence (XRF) to determine the concentrations of Br, Ca, Cu, K, Mn, Pb, Rb, Sr, and A. This technique is widely used because of its accuracy and simplicity of sample preparation, normally complex for this type of biological material. This multi-elemental analysis has proven suitable for wood, a material in which a wide range for each element was encountered in the different species studied: 0.3-5.2 for Br, 126-9074 for Ca, 2.2-11 for Cu, 108-5873 for K, 3.1-134 for Mn, 0.5-4.7 for Pb, 0.3-20 for Rb, 1.2-120 for Sr, and 1.1-20 for Zn (values given in μg g -1 ). (author)

  13. High-throughput screening with micro-x-ray fluorescence

    International Nuclear Information System (INIS)

    Havrilla, George J.; Miller, Thomasin C.

    2005-01-01

    Micro-x-ray fluorescence (MXRF) is a useful characterization tool for high-throughput screening of combinatorial libraries. Due to the increasing threat of use of chemical warfare (CW) agents both in military actions and against civilians by terrorist extremists, there is a strong push to improve existing methods and develop means for the detection of a broad spectrum of CW agents in a minimal amount of time to increase national security. This paper describes a combinatorial high-throughput screening technique for CW receptor discovery to aid in sensor development. MXRF can screen materials for elemental composition at the mesoscale level (tens to hundreds of micrometers). The key aspect of this work is the use of commercial MXRF instrumentation coupled with the inherent heteroatom elements within the target molecules of the combinatorial reaction to provide rapid and specific identification of lead species. The method is demonstrated by screening an 11-mer oligopeptide library for selective binding of the degradation products of the nerve agent VX. The identified oligopeptides can be used as selective molecular receptors for sensor development. The MXRF screening method is nondestructive, requires minimal sample preparation or special tags for analysis, and the screening time depends on the desired sensitivity

  14. Portable x-ray fluorescence spectrometer. Innovative technology summary report

    International Nuclear Information System (INIS)

    1998-12-01

    This report describes the application of portable X-ray fluorescence (XRF) spectrometry to characterize materials related to deactivation and decommissioning (D and D) of contaminated facilities. Two portable XRF instruments manufactured by TN Spectrace were used in a technology evaluation as part of the Large-Scale Demonstration Project (LSDP) held at the Chicago Pile-5 Research Reactor (CP-5) located at Argonne National Laboratory (ANL). The LSDP is sponsored by the US Department of Energy (DOE), Office of Science and Technology, Deactivation and Decommissioning Focus Are (DDFA). The objective of the LSDP is to demonstrate innovative technologies or technology applications potentially beneficial to the D and D of contaminated facilities. The portable XRF technology offers several potential benefits for rapid characterization of facility components and contaminants, including significant cost reduction, fast turnaround time,a nd virtually no secondary waste. Field work for the demonstration of the portable XRF technology was performed from August 28--September 3, 1996 and October 30--December 13, 1996

  15. Non-conventional applications of a noninvasive portable X-ray diffraction/fluorescence instrument

    International Nuclear Information System (INIS)

    Chiari, Giacomo; Sarrazin, Philippe; Heginbotham, Arlen

    2016-01-01

    Noninvasive techniques have become widespread in the cultural heritage analytical domain. The popular handheld X-ray fluorescence (XRF) devices give the elemental composition of all the layers that X-rays can penetrate, but no information on how atoms are bound together or at which depth they are located. A noninvasive portable X-ray powder diffraction/X-ray fluorescence (XRD/XRF) device may offer a solution to these limitations, since it can provide information on the composition of crystalline materials. This paper introduces applications of XRD beyond simple phase recognition. The two fundamental principles for XRD are: (1) the crystallites should be randomly oriented, to ensure proper intensity to all the diffraction peaks, and (2) the material should be positioned exactly in the focal plane of the instrument, respecting its geometry, as any displacement of the sample would results in 2θ shifts of the diffraction peaks. In conventional XRD, the sample is ground and set on the properly positioned sample holder. Using a noninvasive portable instrument, these two requirements are seldom fulfilled. The position, size and orientation of a given crystallite within a layered structure depend on the object itself. Equation correlating the displacement (distance from the focal plane) versus peak shift (angular difference in 2θ from the standard value) is derived and used to determine the depth at which a given substance is located. The quantitative composition of two binary Cu/Zn alloys, simultaneously present, was determined measuring the cell volume and using Vegard's law. The analysis of the whole object gives information on the texture and possible preferred orientations of the crystallites, which influences the peak intensity. This allows for the distinction between clad and electroplated daguerreotypes in the case of silver and between ancient and modern gilding for gold. Analyses of cross sections can be carried out successfully. Finally, beeswax, used in Roman

  16. X-ray fluorescence analysis of praseodymium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Mohile, A.N.

    1976-01-01

    A method for the determination of lanthanum, cerium, neodymium and samarium oxides in praseodymium oxide is described. The sample in the oxalate form is mixed with boric acid binder in the weight ratio of 1:1 and pressed into a pellet. The pellet is irradiated by X-rays from a tungsten tube and fluorescent X-rays are dispersed by a LiF (200) crystal in a Philips semiautomatic X-ray fluorescence spectrometer. The intensity of the characteristic X-rays of the impurity elements is measured by a flow proportional counter at selected 20 angles. The minium determination limit is 0.01% for all impurities. (author)

  17. Romanian medieval earring analysis by X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Therese, Laurent; Guillot, Philippe; Muja, Cristina

    2011-01-01

    Full text: Several instrumental techniques of elemental analysis are now used for the characterization of archaeological materials. The combination between archaeological and analytical information can provide significant knowledge on the constituting material origin, heritage authentication and restoration, provenance, migration, social interaction and exchange. Surface mapping techniques such as X-Ray Fluorescence have become a powerful tool for obtaining qualitative and semi-quantitative information about the chemical composition of cultural heritage materials, including metallic archaeological objects. In this study, the material comes from the Middle Age cemetery of Feldioara (Romania). The excavation of the site located between the evangelical church and the parsonage led to the discovery of several funeral artifacts in 18 graves among a total of 127 excavated. Even if the inventory was quite poor, some of the objects helped in establishing the chronology. Six anonymous Hungarian denarii (silver coins) were attributed to Geza II (1141-1161) and Stefan III (1162-1172), placing the cemetery in the second half of the XII century. This period was also confirmed by three loop shaped earrings with the end in 'S' form (one small and two large earrings). The small earring was found during the excavation in grave number 86, while the two others were discovered together in grave number 113. The anthropological study shown that skeletons excavated from graves 86 and 113 belonged respectively to a child (1 individual, medium level preservation, 9 months +/- 3 months) and to an adult (1 individual). In this work, elemental mapping were obtained by X-ray fluorescence (XRF) technique from Jobin Yvon Horiba XGT-5000 instrument offering detailed elemental images with a spatial resolution of 100μm. The analysis revealed that the earrings were composed of copper, zinc and tin as major elements. Minor elements were also determined. The comparison between the two large earrings

  18. Romanian medieval earring analysis by X-ray fluorescence technique

    Energy Technology Data Exchange (ETDEWEB)

    Therese, Laurent; Guillot, Philippe, E-mail: philippe.guillot@univ-jfc.fr [Laboratoire Diagnostics des Plasmas, CUFR J.F.C, Albi (France); Muja, Cristina [Laboratoire Diagnostics des Plasmas, CUFR J.F.C, Albi (France); Faculty of Biology, University of Bucharest (Romania); Vasile Parvan Institute of Archaeology, Bucharest, (Romania)

    2011-07-01

    Full text: Several instrumental techniques of elemental analysis are now used for the characterization of archaeological materials. The combination between archaeological and analytical information can provide significant knowledge on the constituting material origin, heritage authentication and restoration, provenance, migration, social interaction and exchange. Surface mapping techniques such as X-Ray Fluorescence have become a powerful tool for obtaining qualitative and semi-quantitative information about the chemical composition of cultural heritage materials, including metallic archaeological objects. In this study, the material comes from the Middle Age cemetery of Feldioara (Romania). The excavation of the site located between the evangelical church and the parsonage led to the discovery of several funeral artifacts in 18 graves among a total of 127 excavated. Even if the inventory was quite poor, some of the objects helped in establishing the chronology. Six anonymous Hungarian denarii (silver coins) were attributed to Geza II (1141-1161) and Stefan III (1162-1172), placing the cemetery in the second half of the XII century. This period was also confirmed by three loop shaped earrings with the end in 'S' form (one small and two large earrings). The small earring was found during the excavation in grave number 86, while the two others were discovered together in grave number 113. The anthropological study shown that skeletons excavated from graves 86 and 113 belonged respectively to a child (1 individual, medium level preservation, 9 months +/- 3 months) and to an adult (1 individual). In this work, elemental mapping were obtained by X-ray fluorescence (XRF) technique from Jobin Yvon Horiba XGT-5000 instrument offering detailed elemental images with a spatial resolution of 100{mu}m. The analysis revealed that the earrings were composed of copper, zinc and tin as major elements. Minor elements were also determined. The comparison between the two

  19. Quantum electrodynamics of the internal source x-ray holographies: Bremsstrahlung, fluorescence, and multiple-energy x-ray holography

    International Nuclear Information System (INIS)

    Miller, G.A.; Sorensen, L.B.

    1997-01-01

    Quantum electrodynamics (QED) is used to derive the differential cross sections measured in the three new experimental internal source ensemble x-ray holographies: bremsstrahlung (BXH), fluorescence (XFH), and multiple-energy (MEXH) x-ray holography. The polarization dependence of the BXH cross section is also obtained. For BXH, we study analytically and numerically the possible effects of the virtual photons and electrons which enter QED calculations in summing over the intermediate states. For the low photon and electron energies used in the current experiments, we show that the virtual intermediate states produce only very small effects. This is because the uncertainty principle limits the distance that the virtual particles can propagate to be much shorter than the separation between the regions of high electron density in the adjacent atoms. We also find that using the asymptotic form of the scattering wave function causes about a 5 10% error for near forward scattering. copyright 1997 The American Physical Society

  20. Numerical simulation study for atomic-resolution x-ray fluorescence holography

    International Nuclear Information System (INIS)

    Xie Honglan; Gao Hongyi; Chen Jianwen; Xiong Shisheng; Xu Zhizhan; Wang Junyue; Zhu Peiping; Xian Dingchang

    2003-01-01

    Based on the principle of x-ray fluorescence holography, an iron single crystal model of a body-centred cubic lattice is numerically simulated. From the fluorescence hologram produced numerically, the Fe atomic images were reconstructed. The atomic images of the (001), (100), (010) crystallographic planes were consistent with the corresponding atomic positions of the model. The result indicates that one can obtain internal structure images of single crystals at atomic-resolution by using x-ray fluorescence holography

  1. X-ray film digitization using a personal computer and hand-held scanner: a simple technique for storing images

    International Nuclear Information System (INIS)

    Munoz-Nunez, C. F.; Lloret-Alcaniz, A.

    1998-01-01

    To develop a simple, low-cost technique for the digitization of X-ray films for personal use. A 66-MHz 486 PC with 8 MB of RAM, a Logitech ScanMan 256 hand-held scanner and a standard negatoscope with the power source converted to direct current. Although the system was originally designed for the digitization of mammographies, it has also been used with computed tomography, magnetic resonance, digital angiography and ultrasonographic images, as well as plain X-rays. After a minimal training period, the system digitized X-ray films easily and rapidly. Although the scanning values vary depending on the type of image to be digitized, an input spatial resolution of 200 dpi and a contrast resolution of 256 levels of gray are generally adequate. Of the storage formats tested, JPEG presented the best quality/image size ratio. A simple, low-cost technique has been developed for the digitization of X-ray films. This technique enables the storage of images in a digital format, thus facilitating their presentation and transmission. (Author) 9 refs

  2. X-ray fluorescence analyzers for investigating postmediaeval pottery from Southern Moravia

    International Nuclear Information System (INIS)

    Trojek, Tomas; Hlozek, Matin; Cechak, Tomas; Musilek, Ladislav

    2010-01-01

    This paper deals with an investigation of ceramic archaeological finds with the use of in-situ X-ray fluorescence analysis. Firstly, three configurations of X-ray fluorescence analyzers constructed and used at the Czech Technical University in Prague are described and compared for use in a non-destructive survey of siliceous materials. Detection limits, depth of analysis, the relation of the analyzed area, the homogeneity of the samples, and variations in the element concentrations are discussed. Secondly, many shards of postmediaeval pottery from Southern Moravia are analyzed with X-ray fluorescence analysis and some of them also with electron microprobe analysis. Selected results are described.

  3. A guide for approval of x-ray fluorescence analysis devices

    International Nuclear Information System (INIS)

    1990-01-01

    This guide has been written to assist manufacturers, distributors and users of x-ray fluorescence analysis devices in the preparation of a submission to the Atomic Energy Control Board (AECB) in support of a request for approval of an x-ray fluorescence analysis device. Prior to the issuance of a Radioisotope licence authorizing the use or possession of an x-ray fluorescence analysis device in Canada, the design and construction of the device must be approved by the AECB. The AECB assessment is limited to the radiation safety aspects of use and packaging for transportation

  4. Sampling, storage and sample preparation procedures for X ray fluorescence analysis of environmental materials

    International Nuclear Information System (INIS)

    1997-06-01

    X ray fluorescence (XRF) method is one of the most commonly used nuclear analytical technique because of its multielement and non-destructive character, speed, economy and ease of operation. From the point of view of quality assurance practices, sampling and sample preparation procedures are the most crucial steps in all analytical techniques, (including X ray fluorescence) applied for the analysis of heterogeneous materials. This technical document covers recent modes of the X ray fluorescence method and recent developments in sample preparation techniques for the analysis of environmental materials. Refs, figs, tabs

  5. use of x-ray fluorescence spectrometry to determine trace elements ...

    African Journals Online (AJOL)

    NIJOTECH

    Abstract. This paper deals with application of X-ray fluorescence spectrometry for the detection of trace elements in graphic. An X-ray spectrometer was constructed and used to carry out measurements on graphite spheres impregnated with different chemical elements. The intensities of the lines of these trace elements, ...

  6. Microprocessor-based system for automatic X-ray diffraction and fluorescence

    International Nuclear Information System (INIS)

    Souza, A.M. de; Carmo, L.C.S. do; Pereira, V.J.E.; Soares, E.A.

    1984-01-01

    A data acquisition and processing device appropriate for X-ray analysis and goniometer control was built. The Z-80 based system as well as the whole architeture is described. The advantages and new possibilities of the automated instrument as compared to the traditional ones are listed. The X-ray diffraction and fluorescence techniques can take advantage of the automation. (Author) [pt

  7. Classification and Visualization of Physical and Chemical Properties of Falsified Medicines with Handheld Raman Spectroscopy and X-Ray Computed Tomography.

    Science.gov (United States)

    Kakio, Tomoko; Yoshida, Naoko; Macha, Susan; Moriguchi, Kazunobu; Hiroshima, Takashi; Ikeda, Yukihiro; Tsuboi, Hirohito; Kimura, Kazuko

    2017-09-01

    Analytical methods for the detection of substandard and falsified medical products (SFs) are important for public health and patient safety. Research to understand how the physical and chemical properties of SFs can be most effectively applied to distinguish the SFs from authentic products has not yet been investigated enough. Here, we investigated the usefulness of two analytical methods, handheld Raman spectroscopy (handheld Raman) and X-ray computed tomography (X-ray CT), for detecting SFs among oral solid antihypertensive pharmaceutical products containing candesartan cilexetil as an active pharmaceutical ingredient (API). X-ray CT visualized at least two different types of falsified tablets, one containing many cracks and voids and the other containing aggregates with high electron density, such as from the presence of the heavy elements. Generic products that purported to contain equivalent amounts of API to the authentic products were discriminated from the authentic products by the handheld Raman and the different physical structure on X-ray CT. Approach to investigate both the chemical and physical properties with handheld Raman and X-ray CT, respectively, promise the accurate discrimination of the SFs, even if their visual appearance is similar with authentic products. We present a decision tree for investigating the authenticity of samples purporting to be authentic commercial tablets. Our results indicate that the combination approach of visual observation, handheld Raman and X-ray CT is a powerful strategy for nondestructive discrimination of suspect samples.

  8. X-Ray Fluorescence Spectroscopy for Analysis of Explosive-Related Materials and Unknowns

    Science.gov (United States)

    2017-08-01

    instrument uses a 50 W X-ray tube with a Pd target and has a sample chamber with a six- sample turret. The maximum sample size is 44 mm in diameter by...absorbance of X-rays by the elements present, and the X-ray fluorescence following excitation from other X-rays produced by the sample . Qualitative data...CAN2 Big Approximately 7.6 g of weathered CAN, enough to fill a sample cup To investigate possible errors associated with small sample size

  9. X-ray fluorescence in IAEA Member States: Greece

    International Nuclear Information System (INIS)

    Karydas, A.

    2004-01-01

    The Laboratory of Material Analysis (LMA) of the Institute of Nuclear Physics (INP) at the National Center for Scientific Research (NCSR) 'Demokritos', has been involved very actively during the past few years in the development, evaluation and analytical application of portable X-Ray Fluorescence (XRF) instruments, applied in particular for the non-destructive analysis of cultural materials. The study, conservation and preservation of cultural materials are considered nowadays issues of main concern for countries and international cultural organizations. Due to the strong interest and motivation from archaeologists, conservators and archaeometrical scientists in Greece and elsewhere, a large network has been developed involving the LMA and archaeologists/conservator scientists from Museums (Benaki Museum in Athens), Cultural Foundations (Thera Foundation P. Nomikos), the Greek Ministry of Culture-Conservation Department, Foreign Schools in Greece (American School of Classical Studies, French School of Athens), Universities (Department of Conservation of Antiquities and Works of Art in the Technological Educational Institution of Athens, University of Cincinnatti, Universite de Paris I, Pantheon Sorbonne), private sectors (THETIS, Thetis Authentics - Science and Techniques for Art History Conservation Ltd) and Institutions (Centre de Recherche et de Rastauration des Musees de France, LNS-INFN, LANDIS group). A variety of cultural materials/artifacts have been examined so far, including ceramic vases with colored decoration, bronze artifacts, wall-painting pigments, traces of polychromy on marble sculptures, Gold and Silver ancient jewelry, Gemstones, Roman Coins. Our research and analytical applications of the in-situ XRF analysis have been focused so far on the following: 1) optimum selection and integration of portable XRF instrumentation for improving analytical and sensitivity range; 2) evaluation of the potential of in-situ XRF analysis to provide specific

  10. Characterization of a pulsed x-ray source for fluorescent lifetime measurements

    International Nuclear Information System (INIS)

    Blankespoor, S.C.; Derenzo, S.E.; Moses, W.W.; Rossington, C.S.; Ito, M.; Oba, K.

    1994-01-01

    To search for new, fast, inorganic scintillators, the authors have developed a bench-top pulsed x-ray source for determining fluorescent lifetimes and wavelengths of compounds in crystal or powdered form. This source uses a light-excited x-ray tube which produces x-rays when light from a laser diode strikes its photocathode. The x-ray tube has a tungsten anode, a beryllium exit window, a 30 kV maximum tube bias, and a 50 μA maximum average cathode current. The laser produces 3 x 10 7 photons at 650 nm per ∼100 ps pulse, with up to 10 7 pulses/sec. The time spread for the laser diode, x-ray tube, and a microchannel plate photomultiplier tube is less than 120 ps fwhm. The mean x-ray energy at tube biases of 20, 25, and 30 kV is 9.4, 10.3, and 11.1 keV, respectively. The authors measured 140, 230, and 330 x-ray photons per laser diode pulse per steradian, at tube biases of 20, 25, and 30 kV, respectively. Background x-rays due to dark current occur at a rate of 1 x 10 6 and 3 x 10 6 photons/sec/steradian at biases of 25 and 30 kV, respectively. Data characterizing the x-ray output with an aluminum filter in the x-ray beam are also presented

  11. Application of total reflection X-ray fluorescence spectrometry for ...

    Indian Academy of Sciences (India)

    Applicability of total reflection X-ray fluorescence (TXRF) spectrometry for trace elemental analysis of rainwater samples was studied. The study was used to develop these samples as rainwater standards by the National University of Singapore (NUS). Our laboratory was one of the participants to use TXRF for this study.

  12. Fluorescent x-ray computed tomography to visualize specific material distribution

    Science.gov (United States)

    Takeda, Tohoru; Yuasa, Tetsuya; Hoshino, Atsunori; Akiba, Masahiro; Uchida, Akira; Kazama, Masahiro; Hyodo, Kazuyuki; Dilmanian, F. Avraham; Akatsuka, Takao; Itai, Yuji

    1997-10-01

    Fluorescent x-ray computed tomography (FXCT) is being developed to detect non-radioactive contrast materials in living specimens. The FXCT systems consists of a silicon channel cut monochromator, an x-ray slit and a collimator for detection, a scanning table for the target organ and an x-ray detector for fluorescent x-ray and transmission x-ray. To reduce Compton scattering overlapped on the K(alpha) line, incident monochromatic x-ray was set at 37 keV. At 37 keV Monte Carlo simulation showed almost complete separation between Compton scattering and the K(alpha) line. Actual experiments revealed small contamination of Compton scattering on the K(alpha) line. A clear FXCT image of a phantom was obtained. Using this system the minimal detectable dose of iodine was 30 ng in a volume of 1 mm3, and a linear relationship was demonstrated between photon counts of fluorescent x-rays and the concentration of iodine contrast material. The use of high incident x-ray energy allows an increase in the signal to noise ratio by reducing the Compton scattering on the K(alpha) line.

  13. In vivo x-ray fluorescence of lead and other toxic trace elements

    International Nuclear Information System (INIS)

    Chettle, D.R.

    1995-01-01

    The first in vivo x-ray fluorescence measurements of lead in bone used y-rays from a 57 Co source to excite Pb K x-rays. Later systems used γ-rays from 109 Cd to excite Pb K x-rays or polarized x-rays to excite Ph L x-rays. All three approaches involve an extremely low effective dose to the subject. Of the two K x-ray techniques, 109 Cd is more precise and more flexible in choice of measurement site. Pb L x-ray fluorescence (L-XRF) effectively samples lead at bone surfaces, whereas Ph K x-ray fluorescence (K-XRF) samples through the bulk of a bone. Both the polarized L-XRF and 109 Cd K-XRF achieve similar precision. Renal mercury has recently been determined using a polarized x-ray source, Both renal and hepatic cadmium can be measured using polarized x-rays in conjunction with a Si(Li) detector. Platinum and gold have been measured both by radioisotopic source excitation and by using polarized x-rays, but the latter is to be preferred. Applications of Pb K-XRF have shown that measured bone lead relates strongly to cumulative lead exposure. Secondly, biological half lives of lead in different bone types have been estimated from limited longitudinal data sets and from some cross sectional surveys. Thirdly, the effect of bone lead as an endogenous source of lead has been demonstrated and it has been shown that a majority of circulating blood lead can be mobilized from bone, rather than deriving from new exposure, in some retired lead workers. 35 refs., 5 tabs

  14. Micro-beam X-ray fluorescence and absorption imaging techniques at the IAEA Laboratories

    International Nuclear Information System (INIS)

    Wegrzynek, Dariusz; Markowicz, A.; Bamford, S.; Chinea-Cano, E.; Bogovac, M.

    2005-01-01

    X-ray tube based, micro-beam X-ray fluorescence scanning spectrometer has been equipped with two energy dispersive X-ray detectors. The two-detector configuration allows for simultaneous collection of X-ray fluorescence (XRF) and transmitted X-ray beam signals with a spatial resolution in the range of 10-50 μm, depending on the X-ray focussing element in use. The XRF signal is collected with a standard, liquid nitrogen cooled Si(Li) detector. The X-ray beam transmitted through the sample is acquired with a thermoelectrically cooled, silicon drift (SD) detector. The data acquisition is carried out in a fully automatic way under control of the SPECTOR-LOCATOR software. The software controls the scanning procedure and X-ray spectra acquisition during the scan. The energy dispersive X-ray spectra collected at every 'pixel' are stored for off-line processing. For selected regions of interest (ROI's), the element maps are constructed and displayed on-line. The spectrometer has been used for mapping elemental distributions and for performing 2D- and 3D-tomograpic imaging of minute objects in X-ray absorption and in X-ray fluorescence mode. A unique feature of the described system is simultaneous utilization of the two detectors, Si(Li) and SD, which adds new options for quantitative analysis and data interpretation. Examples of elemental mapping and 3D tomographic imaging as well as the advanced features of the SPECTOR-LOCATOR measurement control and data acquisition software are presented in this work

  15. Simultaneous analysis of gaseous and particulate sulphur in the atmosphere by x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Matsuda, Yatsuka; Mamuro, Tetsuo

    1975-01-01

    An analytical technique for the simultaneous measurements of the atmospheric concentrations of SO 2 gas and sulphur absorbed by aerosol particles has been developed. Aerosol particles are collected on membrane filter and at the same time SO 2 gas is captured on alkali impregnated filter. The sulphur content in each filter is measured by an energy dispersive X -ray fluorescence spectrometer consisting of a Si(Li) semiconductor detector connected to a multi-channel pulse height analyzer and an excitation source of 55 Fe. Two methods are acceptable for the determination of the sulphur content in impregnated filter by X-ray fluorescence analysis. In the first method X-ray fluorescence analysis is made after the collected sulphur gas diffused and distributed uniformly enough throughout the filter, and in the second method X-ray fluorescence analysis gas to be finished before the diffusion of the collected sulphur becomes appreciable. (author)

  16. Sweeping total reflection X-ray fluorescence optimisation to monitor the metallic contamination into IC manufacturing

    International Nuclear Information System (INIS)

    Borde, Yannick; Danel, Adrien; Roche, Agnes; Veillerot, Marc

    2008-01-01

    Among the methods available on the market today to control as metallic contamination in integrated circuit manufacturing, Sweeping Total reflection X-ray Fluorescence mode appears a very good method, providing fast and entire wafer mapping. With the goal of a pertinent use of Sweeping Total reflection X-ray Fluorescence in advanced Integrated Circuit manufacturing this work discusses how acceptable levels of contamination specified by the production (low levels to be detected) can be taken into account. The relation between measurement results (surface coverage, throughput, low limit of detection, limit of quantification, quantification of localized contamination) and Sweeping Total reflection X-ray Fluorescence parameters (number of measurement points and integration time per point) is presented in details. In particular, a model is proposed to explain the mismatch between actual surface contamination in a localized spot on wafer and Total reflection X-ray Fluorescence reading. Both calibration and geometric issues have been taken into account

  17. Forensic application of total reflection X-ray fluorescence spectrometry for elemental characterization of ink samples

    International Nuclear Information System (INIS)

    Dhara, Sangita; Misra, N.L.; Maind, S.D.; Kumar, Sanjukta A.; Chattopadhyay, N.; Aggarwal, S.K.

    2010-01-01

    The possibility of applying Total Reflection X-ray Fluorescence for qualitative and quantitative differentiation of documents printed with rare earth tagged and untagged inks has been explored in this paper. For qualitative differentiation, a very small amount of ink was loosened from the printed documents by smoothly rubbing with a new clean blade without destroying the manuscript. 50 μL of Milli-Q water was put on this loose powder, on the manuscript, and was agitated by sucking and releasing the suspension two to three times with the help of a micropipette. The resultant dispersion was deposited on quartz sample support for Total Reflection X-ray Fluorescence measurements. The Total Reflection X-ray Fluorescence spectrum of tagged and untagged inks could be clearly differentiated. In order to see the applicability of Total Reflection X-ray Fluorescence for quantitative determinations of rare earths and also to countercheck such determinations in ink samples, the amounts of rare earth in painted papers with single rare earth tagged inks were determined by digesting the painted paper in HNO 3 /HClO 4 , mixing this solution with the internal standard and recording their Total Reflection X-ray Fluorescence spectra after calibration of the instrument. The results thus obtained were compared with those obtained by Inductively Coupled Plasma Mass Spectrometry and were found in good agreement. The average precision of the Total Reflection X-ray Fluorescence determinations was 5.5% (1σ) and the average deviation of Total Reflection X-ray Fluorescence determined values with that of Inductively Coupled Plasma Mass Spectrometry was 7.3%. These studies have shown that Total Reflection X-ray Fluorescence offers a promising and potential application in forensic work of this nature.

  18. Determination of silver in ancient coins by X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Pairatana, C.

    1976-01-01

    45 coins of late Ayudhaya and Bangkok periods was analyzed quantitatively by x-ray fluorescence technique using radioisotopic sources Pm - 147/A1 and Am - 241. The fluorescence x-rays were detected by Lithium drifted silicon detector and 1024 channels pulse height analyzer. The procedure was laid a stress on non-destructive methods which could be utilized for analysing various kinds of antiquities and work of arts such as metals, alloys, pottery, ceramics, paper, textile, wood etc

  19. Certification of reference materials by energy-dispersive x-ray fluorescence spectrometry?

    DEFF Research Database (Denmark)

    Christensen, Leif Højslet; Heydorn, Kaj

    1985-01-01

    This paper studies the precision and accuracy that can be achieved using energy-dispersive x-ray fluorescence spectrometry for the determination of total sulphur content in BCR 38 Fly Ash issued by the European Community Bureau of Reference.......This paper studies the precision and accuracy that can be achieved using energy-dispersive x-ray fluorescence spectrometry for the determination of total sulphur content in BCR 38 Fly Ash issued by the European Community Bureau of Reference....

  20. Radionuclide X-ray fluorescence analysis of components of the environment

    International Nuclear Information System (INIS)

    Toelgyessy, J.; Havranek, E.; Dejmkova, E.

    1983-12-01

    The physical foundations and methodology are described of radionuclide X-ray fluorescence analysis. The sources are listed of air, water and soil pollution, and the transfer of impurities into biological materials is described. A detailed description is presented of the sampling of air, soil and biological materials and their preparation for analysis. Greatest attention is devoted to radionuclide X-ray fluorescence analysis of the components of the environment. (ES)

  1. Forensic application of total reflection X-ray fluorescence spectrometry for elemental characterization of ink samples

    Energy Technology Data Exchange (ETDEWEB)

    Dhara, Sangita [Fuel Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India); Misra, N.L., E-mail: nlmisra@barc.gov.i [Fuel Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India); Maind, S.D. [NAA Unit of Central Forensic Science Laboratory Hyderabad at Analytical Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India); Kumar, Sanjukta A. [Analytical Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India); Chattopadhyay, N. [NAA Unit of Central Forensic Science Laboratory Hyderabad at Analytical Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India); Aggarwal, S.K. [Fuel Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 085 (India)

    2010-02-15

    The possibility of applying Total Reflection X-ray Fluorescence for qualitative and quantitative differentiation of documents printed with rare earth tagged and untagged inks has been explored in this paper. For qualitative differentiation, a very small amount of ink was loosened from the printed documents by smoothly rubbing with a new clean blade without destroying the manuscript. 50 muL of Milli-Q water was put on this loose powder, on the manuscript, and was agitated by sucking and releasing the suspension two to three times with the help of a micropipette. The resultant dispersion was deposited on quartz sample support for Total Reflection X-ray Fluorescence measurements. The Total Reflection X-ray Fluorescence spectrum of tagged and untagged inks could be clearly differentiated. In order to see the applicability of Total Reflection X-ray Fluorescence for quantitative determinations of rare earths and also to countercheck such determinations in ink samples, the amounts of rare earth in painted papers with single rare earth tagged inks were determined by digesting the painted paper in HNO{sub 3}/HClO{sub 4}, mixing this solution with the internal standard and recording their Total Reflection X-ray Fluorescence spectra after calibration of the instrument. The results thus obtained were compared with those obtained by Inductively Coupled Plasma Mass Spectrometry and were found in good agreement. The average precision of the Total Reflection X-ray Fluorescence determinations was 5.5% (1sigma) and the average deviation of Total Reflection X-ray Fluorescence determined values with that of Inductively Coupled Plasma Mass Spectrometry was 7.3%. These studies have shown that Total Reflection X-ray Fluorescence offers a promising and potential application in forensic work of this nature.

  2. ISS Ammonia Leak Detection Through X-Ray Fluorescence

    Science.gov (United States)

    Camp, Jordan; Barthelmy, Scott; Skinner, Gerry

    2013-01-01

    Ammonia leaks are a significant concern for the International Space Station (ISS). The ISS has external transport lines that direct liquid ammonia to radiator panels where the ammonia is cooled and then brought back to thermal control units. These transport lines and radiator panels are subject to stress from micrometeorites and temperature variations, and have developed small leaks. The ISS can accommodate these leaks at their present rate, but if the rate increased by a factor of ten, it could potentially deplete the ammonia supply and impact the proper functioning of the ISS thermal control system, causing a serious safety risk. A proposed ISS astrophysics instrument, the Lobster X-Ray Monitor, can be used to detect and localize ISS ammonia leaks. Based on the optical design of the eye of its namesake crustacean, the Lobster detector gives simultaneously large field of view and good position resolution. The leak detection principle is that the nitrogen in the leaking ammonia will be ionized by X-rays from the Sun, and then emit its own characteristic Xray signal. The Lobster instrument, nominally facing zenith for its astrophysics observations, can be periodically pointed towards the ISS radiator panels and some sections of the transport lines to detect and localize the characteristic X-rays from the ammonia leaks. Another possibility is to use the ISS robot arm to grab the Lobster instrument and scan it across the transport lines and radiator panels. In this case the leak detection can be made more sensitive by including a focused 100-microampere electron beam to stimulate X-ray emission from the leaking nitrogen. Laboratory studies have shown that either approach can be used to locate ammonia leaks at the level of 0.1 kg/day, a threshold rate of concern for the ISS. The Lobster instrument uses two main components: (1) a microchannel plate optic (also known as a Lobster optic) that focuses the X-rays and directs them to the focal plane, and (2) a CCD (charge

  3. Method for detecting binding events using micro-X-ray fluorescence spectrometry

    Science.gov (United States)

    Warner, Benjamin P.; Havrilla, George J.; Mann, Grace

    2010-12-28

    Method for detecting binding events using micro-X-ray fluorescence spectrometry. Receptors are exposed to at least one potential binder and arrayed on a substrate support. Each member of the array is exposed to X-ray radiation. The magnitude of a detectable X-ray fluorescence signal for at least one element can be used to determine whether a binding event between a binder and a receptor has occurred, and can provide information related to the extent of binding between the binder and receptor.

  4. LCLS in—photon out: fluorescence measurement of neon using soft x-rays

    Science.gov (United States)

    Obaid, Razib; Buth, Christian; Dakovski, Georgi L.; Beerwerth, Randolf; Holmes, Michael; Aldrich, Jeff; Lin, Ming-Fu; Minitti, Michael; Osipov, Timur; Schlotter, William; Cederbaum, Lorenz S.; Fritzsche, Stephan; Berrah, Nora

    2018-02-01

    We measured the fluorescence photon yield of neon upon soft x-ray ionization (∼1200 eV) from the x-ray free-electron laser at Linac Coherent Light Source, and demonstrated the usage of a grazing incidence spectrometer with a variable line spacing grating to perform x-ray fluorescence spectroscopy on a gas phase system. Our measurements also allowed us to estimate the focal size of the beam from the theoretical description developed, in terms of the rate equation approximation accounting for photoionization shake off of neutral neon and double auger decay of single core holes.

  5. The application of X-ray fluorescence spectrometry to prospecting potential gold deposits

    International Nuclear Information System (INIS)

    Shang Fengjun; Wang Haixia; Zhou Rongsheng

    2001-01-01

    The fieldwork high-sensitivity X-ray fluorescence analysis (FXFA) adopting miniaturized X-ray tube, Si-PIN detector with peltier cooler and notebook PC spectrometry is presented. Using this system, the authors carried out a preliminary research of its application to some gold mine in Sichuan. According to the close relationship between the high-grade element arsenic and gold in ore-forming components, X-ray fluorescence spectrometry can be used to reveal the existence of potential gold mineralization in fields rapidly. This is of great significance in guiding the field geological collection

  6. Multiple X-ray tomography using transmitted, scattered and fluorescent radiation

    International Nuclear Information System (INIS)

    Cesareo, R.; Brunetti, A.; Golosio, B.; Lopes, R.T.; Barroso, R.C.; Donativi, M.; Castellano, A.; Quarta, S.

    2003-01-01

    A multiple CT-scanner is described, which contemporaneously uses transmitted, scattered and fluorescent X-rays for Imaging. The scanner is characterized by a small size X-ray tube and by four detectors: a ''pencil'' X-ray NaI(Tl) for transmitted tomography, a larger size NaI(Tl) for 90 C o Compton tomography, a thermoelectrically cooled Si-PIN or CdZnTe for fluorescent imaging and a CdZnTe for Rayleigh (or diffraction) tomography. Examples of applications are shown

  7. X-ray fluorescence holography: A different approach to data collection

    International Nuclear Information System (INIS)

    Busetto, E.; Kopecky, M.; Lausi, A.; Menk, R.H.; Miculin, M.; Savoia, A.

    2000-01-01

    The images of nearest neighbors of gallium atoms in a GaAs crystal were obtained by the x-ray fluorescence holography technique. The fluorescence from gallium atoms was selected by means of a thin zinc foil filter that made possible the use of an x-ray silicon photodiode detector without energy resolution. This method makes possible the detection of a much higher signal with respect to all previous experiments, thus reducing drastically measuring times, that is a basic and essential step from contemporary demonstration experiments to possible practical applications of x-ray holography in structure analysis

  8. X-ray fluorescence in Member States (Italy): Full field X-ray fluorescence imaging with high-energy and high-spatial resolution

    Energy Technology Data Exchange (ETDEWEB)

    Romano, F. P.; Masini, N.; Pappalardo, L., E-mail: romanop@lns.infn.it [IBAM, CNR, Via Biblioteca 4, 95124 Catania (Italy); Cosentino, L.; Gammino, S.; Mascali, D.; Rizzo, F. [INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy)

    2014-02-15

    A full field X-ray camera for the X-Ray Fluorescence imaging of materials with high-energy and high-spatial resolution was designed and developed. The system was realized by coupling a pinhole collimator with a positionsensitive CCD detector. X-Ray fluorescence is induced on the samples by irradiation with an external X-ray tube. The characteristic X-ray spectra of the investigated materials are obtained by using a multi-frames acquisition in single-photon counting. The energy resolution measured at the Fe-Kα line was 157 eV. The spatial resolution of the system was determined by the analysis of a sharp-edge at different magnification values; it was estimated to be 90 μm at a magnification value of 3.2x and 190 μm at 0.8x. The present set-up of the system is suited to analyze samples with dimensions up to 5x4 cm{sup 2}. Typical measurement time is in the range between 1h to 4 h. (author)

  9. Maintaining radiation exposures as low as reasonably achievable (ALARA) for dental personnel operating portable hand-held x-ray equipment.

    Science.gov (United States)

    McGiff, Thomas J; Danforth, Robert A; Herschaft, Edward E

    2012-08-01

    Clinical experience indicates that newly available portable hand-held x-ray units provide advantages compared to traditional fixed properly installed and operated x-ray units in dental radiography. However, concern that hand-held x-ray units produce higher operator doses than fixed x-ray units has caused regulatory agencies to mandate requirements for use of hand-held units that go beyond those recommended by the manufacturer and can discourage the use of this technology. To assess the need for additional requirements, a hand-held x-ray unit and a pair of manikins were used to measure the dose to a simulated operator under two conditions: exposures made according to the manufacturer's recommendations and exposures made according to manufacturer's recommendation except for the removal of the x-ray unit's protective backscatter shield. Dose to the simulated operator was determined using an array of personal dosimeters and a pair of pressurized ion chambers. The results indicate that the dose to an operator of this equipment will be less than 0.6 mSv y⁻¹ if the device is used according to the manufacturer's recommendations. This suggests that doses to properly trained operators of well-designed, hand-held dental x-ray units will be below 1.0 mSv y⁻¹ (2% of the annual occupational dose limit) even if additional no additional operational requirements are established by regulatory agencies. This level of annual dose is similar to those reported as typical dental personnel using fixed x-ray units and appears to satisfy the ALARA principal for this class of occupational exposures.

  10. Scoria analysis by X-ray fluorescence with briquet

    International Nuclear Information System (INIS)

    Silva, G.R. da; Albert, P.E.

    1987-01-01

    The sample of scoria is triturated, using boric acid as agglomerative. The briquets are analyzed by X-ray quantometry, establishing the proportions of SiO 2 , P 2 O 5 , S, Cr 2 O 3 , V 2 O 5 , TiO 2 , Al 2 O 3 , Mg O, Ca O, FeO and their basicity. The calibration curves are prepared with standard scoria samples, by chemical analysis. (C.G.C.) [pt

  11. Development of a fluorescent x-ray source for medical imaging

    Science.gov (United States)

    Toyofuku, F.; Tokumori, K.; Nishimura, K.; Saito, T.; Takeda, T.; Itai, Y.; Hyodo, K.; Ando, M.; Endo, M.; Naito, H.; Uyama, C.

    1995-02-01

    A fluorescent x-ray source for medical imaging, such as K-edge subtraction angiography and monochromatic x-ray CT, has been developed. Using a 6.5 GeV accumulation ring in Tsukuba, fluorescent x rays, which range from about 30 to 70 keV are generated by irradiating several target materials. Measurements have been made of output intensities and energy spectra for different target angles and extraction angles. The intensities of fluorescent x rays at a 30 mA beam current are on the order of 1-3×106 photons/mm2/s at 30 cm from the local spot where the incident beam is collimated to 1 mm2. A phantom which contains three different contrast media (iodine, barium, gadolinium) was used for the K-edge energy subtraction, and element selective CT images were obtained.

  12. X-ray fluorescence analysis of thin films at glancing-incident and -takeoff angles

    International Nuclear Information System (INIS)

    Tsuji, K.; Sato, S.; Hirokawa, K.

    1995-01-01

    We have developed a new analytical method, Glancing-Incidence and -Takeoff X-Ray Fluorescence (GIT-XRF) method for the first time. Here, we present an idea for a thin-film analysis and a surface analysis by the GIT-XRF method. In this method, the dependence of the fluorescent x-ray intensity on takeoff angle is measured at various incident angles of the primary x-ray. Compared with a total reflection x-ray fluorescence method, the GIT-XRF method allows a detailed thin-film analysis, because the thin film is cross-checked by many experimental curves. Moreover, a surface-sensitive analysis is also possible by the GIT-XRF method. (author)

  13. Elemental Composition of Mars Return Samples Using X-Ray Fluorescence Imaging at the National Synchrotron Light Source II

    Science.gov (United States)

    Thieme, J.; Hurowitz, J. A.; Schoonen, M. A.; Fogelqvist, E.; Gregerson, J.; Farley, K. A.; Sherman, S.; Hill, J.

    2018-04-01

    NSLS-II at BNL provides a unique and critical capability to perform assessments of the elemental composition and the chemical state of Mars returned samples using synchrotron radiation X-ray fluorescence imaging and X-ray absorption spectroscopy.

  14. Characterization of a confocal three-dimensional micro X-ray fluorescence facility based on polycapillary X-ray optics and Kirkpatrick-Baez mirrors

    International Nuclear Information System (INIS)

    Sun Tianxi; Ding Xunliang; Liu Zhiguo; Zhu Guanghua; Li Yude; Wei Xiangjun; Chen Dongliang; Xu Qing; Liu Quanru; Huang Yuying; Lin Xiaoyan; Sun Hongbo

    2008-01-01

    A new confocal three-dimensional micro X-ray fluorescence (3D micro-XRF) facility based on polycapillary X-ray optics in the detection channel and Kirkpatrick-Baez (KB) mirrors in the excitation channel is designed. The lateral resolution (l x , l y ) of this confocal three-dimensional micro-X-ray fluorescence facility is 76.3(l x ) and 53.4(l y ) μm respectively, and its depth resolution d z is 77.1 μm at θ = 90 o . A plant sample (twig of B. microphylla) and airborne particles are analyzed

  15. Development of a portable system of X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Mantuano, Andrea; Crisostomo, Jose V.V.; Barros, Mariana J.; Oliveira, Luis F.; Barroso, Regina C.

    2009-01-01

    This paper develops a compact and portable spectrometry system that will be used at the Laboratory of Applied Physics to the Biomedical and Environmental Sciences of the Institute of Physics/UERJ, Rio de Janeiro, Brazil. The laboratory both prepares the samples and develops the X-ray spectrometry techniques. The techniques of X-ray diffraction and fluorescence on various samples (biological, industrial and environmental) are used, attending to pos-graduation and graduation students, with multidisciplinary characteristics. The Mini-X system consists of X-ray mini tube MINI-X from Amptek with tungsten (W) target, and a compact spectrometer X123, also from Amptek that includes a detector, pre-amplifier, digital pulse processor, and multichannel. All the system is controlled by dedicated microprocessor. This work will present both a methodology for alignment and calibration of the system as far the first measurements performed using the X-ray fluorescence technique on standard samples. The multi elementary analysis by X-ray fluorescence (XRF) is based on the measurements of the characteristic X-ray intensity emitted by the chemical elements components of the samples when excited. Therefore, from the development of this compact and versatile system it will be possible to obtain the fluorescent intensities of the analysed samples at the Laboratory, not only at the research area but at the teaching area. Besides, new laboratory practices are being developed for the discipline of medical physics

  16. Characteristics of X-ray fluorescence of nuclear materials

    Energy Technology Data Exchange (ETDEWEB)

    Park, Seunghoon; Kwak, Sung-Woo; Shin, Jung-Ki; Park, Uk-Rayng; Jung, Heejun [Korea Institute of Nuclear Nonproliferation and Control, Daejeon (Korea, Republic of)

    2015-10-15

    LED is a technique of determination of uranium concentration as a continuous X-ray energy beams transmit a uranium liquid sample for safeguards. Compared to K-edge densitometer, due to relatively lower energy (L-edge energy is 17.17 keV) of Uranium L series energy than K-series energy, L-edge densitometer does not require high purity germanium detector with liquid nitride cooling. Therefore, the Ledge densitometer is appropriate for portable equipment for on-site nuclear material inspection and safeguards at facility sites. XRF combined with LED is a technique of finding of nuclear materials from reflected characteristic X-ray photons. In this study, characteristics of XRF of nuclear materials are simulated Monte Carlo method (Geant4) for feasibility of the system for determination of concentration of nuclear species. The analysis method of uranium concentration or minor actinides is applied using combination of linear extrapolation from jump of L-edge of sample and ratio between uranium and minor actinide from XRF measurement. In this study, The XRF ch aracteristics was simulated from Monte Carlo method. The peaks were obtained from nuclear material mixture. The estimated nuclear material concentration is low due to the volume effect of the sample. The correction factor or minimization of the effect is required.

  17. Automatic energy dispersive x-ray fluorescence analysing apparatus

    International Nuclear Information System (INIS)

    Russ, J.C.; Carey, R.; Chopra, V.K.

    1983-01-01

    The invention discloses a number of improvements for an energy dispersive X-ray analysis system having computer supervised data collection, display and processing. The systems with which the improved circuitry and methods may be used include a dual interlocking bus structure so that the analyzer and computer functions communicate directly with each other and the user has immediate keyboard control of both. Such a system normally includes a system base control, a control console and a display console. The portions of the system which have been improved include a new type of ratemeter which gives a voltage output proportional to the intensity of the energy window or windows under consideration, an output which is an absolute digital representation of the intensity count rate, circuitry for input multiplexing and multiple output voltage buffering of the ratemeter to accomodate multiple single channel signals, and a new dead time correction to enable meaningful single channel intensity data to be handled by the system. An extension of the ratemeter is also disclosed for use in conjunction with X-ray mapping, enabling enhancements to be made on mapping SCA data

  18. X-ray Fluorescence Spectroscopy: the Potential of Astrophysics-developed Techniques

    Science.gov (United States)

    Elvis, M.; Allen, B.; Hong, J.; Grindlay, J.; Kraft, R.; Binzel, R. P.; Masterton, R.

    2012-12-01

    X-ray fluorescence from the surface of airless bodies has been studied since the Apollo X-ray fluorescence experiment mapped parts of the lunar surface in 1971-1972. That experiment used a collimated proportional counter with a resolving power of ~1 and a beam size of ~1degree. Filters separated only Mg, Al and SI lines. We review progress in X-ray detectors and imaging for astrophysics and show how these advances enable much more powerful use of X-ray fluorescence for the study of airless bodies. Astrophysics X-ray instrumentation has developed enormously since 1972. Low noise, high quantum efficiency, X-ray CCDs have flown on ASCA, XMM-Newton, the Chandra X-ray Observatory, Swift and Suzaku, and are the workhorses of X-ray astronomy. They normally span 0.5 to ~8 keV with an energy resolution of ~100 eV. New developments in silicon based detectors, especially individual pixel addressable devices, such as CMOS detectors, can withstand many orders of magnitude more radiation than conventional CCDs before degradation. The capability of high read rates provides dynamic range and temporal resolution. Additionally, the rapid read rates minimize shot noise from thermal dark current and optical light. CMOS detectors can therefore run at warmer temperatures and with ultra-thin optical blocking filters. Thin OBFs mean near unity quantum efficiency below 1 keV, thus maximizing response at the C and O lines.such as CMOS detectors, promise advances. X-ray imaging has advanced similarly far. Two types of imager are now available: specular reflection and coded apertures. X-ray mirrors have been flown on the Einstein Observatory, XMM-Newton, Chandra and others. However, as X-ray reflection only occurs at small (~1degree) incidence angles, which then requires long focal lengths (meters), mirrors are not usually practical for planetary missions. Moreover the field of view of X-ray mirrors is comparable to the incident angle, so can only image relatively small regions. More useful

  19. Synchrotron X-ray fluorescence analysis in environmental and earth sciences

    Directory of Open Access Journals (Sweden)

    Adams F.

    2010-12-01

    Full Text Available Compared to other microscopic analytical tools X-ray microscopy techniques have the advantage that the large penetration depth of X-rays in matter allows one to investigate the interior of an object without destructive sample preparation. In combination with X-ray fluorescence tomography, analytical information from inside of a specimen can be obtained. Different X-ray analytical techniques can be used to produce contrast, X-ray absorption, fluorescence, and diffraction, to yield chemical, elemental, and structural information about the sample. Scanning microscopy on the basis of various lens systems in synchrotron radiation sources provides a routine spatial resolution of now about 100 nanometer but in the foreseeable future a 10–20 nanometer spatial resolution can be expected. X-ray absorption spectrometry can also provide chemical (speciation information on the sample. All this makes X-ray microscopy attractive to many fields of science. In this paper the techniques are briefly reviewed and a number of applications in the earth, planetary and cosmos sciences are illustrated with state-of-the art examples, while applications in the environmental sciences and biology are also briefly discussed.

  20. X-ray fluorescence analysis of neodymium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Mohile, A.N.

    1977-01-01

    An X-ray fluorescence method for the determination of cesium, praseodymium, samarium, europium and gadolinium in pure neodymium oxide and oxalate is described. The oxide sample is converted to oxalate and mixed with a binder (boric acid) to obtain a pressed circular pellet. The amount of sample needed for analysis is reduced by making use of the double layer pellet technique. A tungsten target X-ray tube is employed to irradiate the sample and a Philips PW 1220 semiautomatic X-ray spectrometer with a LiF (200) crystal is used to analyse the fluorescent X-rays. The minimum determination limit is 0.01 percent for all rare earths determined except for europium for which the limit is 0.005 percent. Three sigma detection limits have been calculated. (author)

  1. Fundamental parameters method for quantitative energy dispersive x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Demirel, H.; Zararsiz, A.

    1986-01-01

    In this study, the requirement of the standart material in photon excited energy distributed X-ray fluorescence analysis has been removed. The interaction of X-rays with matter has been taken into account. A computer program has been developed by using the fundamental parameters of X-ray fluorescence technique and the spectral intensity 'K' of pure elements at saturation thickness has been obtained. For experimental purpose a convenient source-target-detector geometry has been designed. In order to excite the samples,Cd-109 radioisotope source has been used. The peak intensities has been obtained in a vacum chamber by counting the emitted X-rays. The calculation of concentration has been performed for double mixed samples correcting the effects of absorption and enchancement factors. The results were in conformity with their certificate values. (author)

  2. Integrated image presentation of transmission and fluorescent X-ray CT using synchrotron radiation

    Energy Technology Data Exchange (ETDEWEB)

    Zeniya, T.; Takeda, T. E-mail: ttakeda@md.tsukuba.ac.jp; Yu, Q.; Hasegawa, Y.; Hyodo, K.; Yuasa, T.; Hiranaka, Y.; Itai, Y.; Akatsuka, T

    2001-07-21

    We have developed a computed tomography (CT) system with synchrotron radiation (SR) to detect fluorescent X-rays and transmitted X-rays simultaneously. Both SR transmission X-ray CT (SR-TXCT) and SR fluorescent X-ray CT (SR-FXCT) can describe cross-sectional images with high spatial and contrast resolutions as compared to conventional CT. TXCT gives morphological information and FXCT gives functional information of organs. So, superposed display system for SR-FXCT and SR-TXCT images has been developed for clinical diagnosis with higher reliability. Preliminary experiment with brain phantom was carried out and the superposition of both images was performed. The superposed SR-CT image gave us both functional and morphological information easily with high reliability, thus demonstrating the usefulness of this system.

  3. Integrated image presentation of transmission and fluorescent X-ray CT using synchrotron radiation

    Science.gov (United States)

    Zeniya, T.; Takeda, T.; Yu, Q.; Hasegawa, Y.; Hyodo, K.; Yuasa, T.; Hiranaka, Y.; Itai, Y.; Akatsuka, T.

    2001-07-01

    We have developed a computed tomography (CT) system with synchrotron radiation (SR) to detect fluorescent X-rays and transmitted X-rays simultaneously. Both SR transmission X-ray CT (SR-TXCT) and SR fluorescent X-ray CT (SR-FXCT) can describe cross-sectional images with high spatial and contrast resolutions as compared to conventional CT. TXCT gives morphological information and FXCT gives functional information of organs. So, superposed display system for SR-FXCT and SR-TXCT images has been developed for clinical diagnosis with higher reliability. Preliminary experiment with brain phantom was carried out and the superposition of both images was performed. The superposed SR-CT image gave us both functional and morphological information easily with high reliability, thus demonstrating the usefulness of this system.

  4. X-ray fluorescence analysis of erbium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Khanna, P.P.; Thomas, A.

    1981-01-01

    A method for the determination of Tb, Dy, Ho, Tm, Yb, Lu and Y oxides in Er 2 O 3 is described. 450 mg sample in the oxalate form is mixed with 150 mg boric acid binding material and pressed into a 1.25 inch diameter pellet over a supporting pellet of boric acid. The sample is then irradiated by X-rays from a tungsten tube and the fluorescent X-rays are dispersed by a LiF (200) crystal in a Philips PW 1220 semiautomatic X-ray fluorescence spectrometer. The intensities of characteristic X-rays of the impurity elements are measured by a flow proportional counter for all elements except yttrium for which the intensities are measured by a scintillation counter. The lowest determination limit is 0.005% for all impurities except for Yb for which it is 0.01%. Calculations for theoretical detection limit are given. (author)

  5. X-ray fluorescence analysis of ytterbium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Khanna, P.P.; Thomas, A.

    1982-01-01

    An XRF method for the determination of Ho, Er, Tm, Lu and Y oxides in Yb 2 O 3 is described. 450 mg sample in the oxalate form is mixed with 150 mg boric acid binding material and pressed into 1.25 inch diameter pellet over a supporting pellet of boric acid. The sample is irradiated by X-rays from a tungsten tube and the fluorescent X-rays are dispersed by a LiF(200) crystal in a Philips PW 1220 semiautomatic X-ray fluorescence spectrometer. The intensities of characteristic X-rays of the impurity elements are measured by a flow proportional counter or a scintillation counter. The lowest determination limit is 0.005% for Ho, Er, Tm and Y and 0.01% for Lu. Calculations for theoretical detection limit, standard deviation and uncertainty are done and presented. (author)

  6. X-ray fluorescence hologram data collection with a cooled avalanche photodiode

    CERN Document Server

    Hayashi, K; Matsubara, E I; Kishimoto, S; Mori, T; Tanaka, M

    2002-01-01

    A high counting rate X-ray detector with an appropriate energy resolution is desired for high quality X-ray fluorescence hologram measurements because a holographic pattern is detected as extremely small intensity variations of X-ray fluorescence on a large intensity background. A cooled avalanche photodiode (APD), which has about 10% energy resolution and is designed for a high counting rate, fits the above requirements. Reconstructed atomic images from experimental holograms using the APD system provide us a clear view of the first and second neighbor atoms around an emitter. The present result proved that a combination of this APD system and a synchrotron X-ray source enables us to measure a high quality hologram for a reasonable measurement time.

  7. Viability of exploiting L-shell fluorescence for X-ray polarimetry

    Energy Technology Data Exchange (ETDEWEB)

    Weisskopf, M C; Elsner, R F; Ramsey, B D [National Aeronautics and Space Administration, Huntsville, AL (USA). Space Sciences Lab.; Sutherland, P G [McMaster Univ., Hamilton, Ontario (Canada). Dept. of Physics

    1985-05-15

    It has been suggested that one may build an X-ray polarimeter by exploiting the polarization dependence of the angular distribution of L-shell fluorescence photons. In this paper we examine, theoretically, the sensitivity of this approach to polarimetry. We apply our calculations to several detection schemes using imaging proportional counters that would have direct application in X-ray astronomy. We find, however, that the sensitivity of this method for measuring X-ray polarization is too low to be of use for other than laboratory applications.

  8. On the viability of exploiting L-shell fluorescence for X-ray polarimetry

    Energy Technology Data Exchange (ETDEWEB)

    Weisskopf, M C; Elsner, R F; Ramsey, B D; Sutherland, P G

    1985-05-15

    It has been suggested that one may build an X-ray polarimeter by exploiting the polarization dependence of the angular distribution of L-shell fluorescence photons. In this paper we examine, theoretically, the sensitivity of this approach to polarimetry. We apply our calculations to several detection schemes using imaging proportional counters that would have direct application in X-ray astronomy. We find, however, that the sensitivity of this method for measuring X-ray polarization is too low to be of use for other than laboratory applications. (orig.).

  9. Elemental analysis of air particulate samples in Jakarta area by x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Yumiarti; Yusuf, M.; Mellawati, June; Menry, Yulizon; Surtipanti S

    1998-01-01

    Determination of elements in air particulate samples collected from Jakarta, especially from industrial area Pulo Gadung, also from residence, office, and recreation sites had been carried out. The samples collected periodically from August through December 1996. The elements were analyzed by X-ray fluorescence spectrometry method. Quantitative and qualitative analyses were done using QXAS AXIL (Quantitative X-ray Analysis System of x-ray Spectra by Iterative Least squares fitting) and QAES (Quantitative Analyses of Environmental Samples) package program. Results of the analyses showed that the content of heavy metal elements in air particulate samples from all areas studied were still below the maximum permissible concentration. (authors)

  10. Quantitative schemes in energy dispersive X-ray fluorescence implemented in AXIL

    International Nuclear Information System (INIS)

    Tchantchane, A.; Benamar, M.A.; Tobbeche, S.

    1995-01-01

    E.D.X.R.F (Energy Dispersive X-ray Fluorescence) has long been used for quantitative analysis of many types of samples including environment samples. the software package AXIL (Analysis of x-ray spectra by iterative least quares) is extensively used for the spectra analysis and the quantification of x-ray spectra. It includes several methods of quantitative schemes for evaluating element concentrations. We present the general theory behind each scheme implemented into the software package. The spectra of the performance of each of these quantitative schemes. We have also investigated their performance relative to the uncertainties in the experimental parameters and sample description

  11. Energy dispersion X-ray fluorescence techniques in water pollution analysis

    International Nuclear Information System (INIS)

    Holynska, B.

    1980-01-01

    Advantages and limitations of energy dispersion X-ray fluorescence methods for analysis of pollutants in water are discussed. The necessary equipment for X-ray measurement of insoluble and dissolved trace metals in water is described. Different techniques of enrichment of trace metals are presented: ion exchange on selective Chelex-100 exchanger, precipitation with chelating agents DDTC and APDC, and adsorption on activated carbon. Some results obtained using different preconcentration methods for trace metals determination in different waters are presented. (author)

  12. Correction for interelement effect in X-Ray fluorescence analysis of trace elements in geological materials

    International Nuclear Information System (INIS)

    El-Behay, A.Z.; Attawiya, M.Y.; Khattab, F.M.

    1984-01-01

    In a trial to obtain accurate results from X-ray fluorescence technique for the analysis of trace elements in geological materials, two corrections were used for the obtained data, namely, correction for the observed x-ray intensities for absorption and/or enhancement effects due to the presence of other elements in the system and correction for spectral deconvolution to account for the overlapping lines. Significant improvement in the precision and accuracy was obtained and evaluated

  13. Application of the nuclear x-ray fluorescence method to prospecting for gold in-situ

    International Nuclear Information System (INIS)

    Zhang, Y.; Xie, T.; Zhou, S.; Ge, L.

    1989-01-01

    Arsenic and chalcophile elements are often associated with gold, and can be considered indicator elements when prospecting for gold deposits. The nuclear geophysics X-ray fluorescence method can be used to search for hidden gold deposits by measuring fluorescence intensities of the indicator elements in situ. The method can speed geologic investigation and reduce exploration cost. Three types of portable radioisotope X-ray fluorescence analyzers, designed and manufactured by Chengdu College of Geology and Chongqing Geological Instrument Factory, are briefly introduced. These analyzers are widely used in different stages of geologic investigation for gold in China. In the two case histories presented five anomalous zones of X-ray fluorescence intensity related to gold mineralization are located and one hidden gold deposit is discovered with gold content of 23 g/t

  14. Disparity in formulations used for fluorescent X-ray intensity measurements

    International Nuclear Information System (INIS)

    Mittal, Raj; Gupta, Sheenu

    2011-01-01

    The paper presents a problem in computations of X-ray fluorescence cross-sections, shell/sub-shell fluorescence yields, Coster-Kronig yields, vacancy alignment, etc. from X-ray fluorescence (XRF) studies. While using barn/atom as a unit for cross-sections if the atomic masses are not considered it causes a discrepancy in the measured cross-section, yield and alignment values. Most of the earlier publications are being quoted where such an oversight has occurred and discrepancy is evident. - Highlights: → Manuscript gives basic formulation for measurements of fluorescent X-ray intensities. → Most publications ignored the fact that use of barn/atom units for cross-sections requires atomic masses. → Published experimental results higher by a factor ≥2 or less by factor 1/M K . → Inspection of published data on XRF parameters needed.

  15. Determination of chlorine in coal by X-ray fluorescence spectrometry method

    Energy Technology Data Exchange (ETDEWEB)

    Marek, S.; Bojarska, K. [Central Mining Institute, Katowice (Poland). Dept. of Environmental Monitoring

    1997-12-31

    Determination of chlorine contents in coal is essential for both environmental protection and its technological use. The existing method of chlorine determination in coal are titration methods which have considerable errors particularly in the low concentration range. The elaborated method with the use of X-ray fluorescence spectrometry in a comparison to the other methods is much faster and has better precision and accuracy. The principle of the method lies in the measurement of X-ray fluorescence radiation intensity which is emitted by chlorine in a sample and its comparison with standards. The calibration of the elaborated XRF method is based on natural coals having various concentrations of chlorine within the whole range of its occurrence in Polish coals. Concentrations for the calibration purpose were obtained by the determination of chlorine contents in selected coals by atomic absorption spectrometry method. The procedure of sample preparation for direct X-ray measurements, instrumental measuring conditions and the way of calibration curve preparation are described in the paper. All X-ray measurements were done with a Phillips sequential X-ray fluorescence spectrometer. A double anode Cr-Au X-ray tube with maximum power 3000 MW was used as the excitation source. 5 figs., 4 tabs.

  16. Assembly of positioner of automated two-dimensional scan coupled to X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Silva, Leonardo Santiago Melgaço

    2011-01-01

    This work describes the design and assembling of a prototype automated positioner two-dimensional scanning coupled to X-ray fluorescence spectrometry. The work aims to achieve a portable and easy to use, device of broad utility in the analysis of samples by X-ray fluorescence area of expertise and research. The two-dimensional scanning of the positioner is by means of two stepper motors controlled by a microcontroller PIC 16F877A, encoder and optical sensors. The user interacts with the XY table through an interface program for the Windows operating system, which communicates with the microcontroller through the serial port. The system of Fluorescence Spectroscopy incorporated into the positioner consists of a system commercially available system from the company AMPTEK, where the primary source of excitation of the sample was a source of 241 Am of 59.5 KeV emissions. Resolution and accuracy of tests were performed in the XY scanning process and reproducibility of the same kit with the fluorescence spectrometry X-ray. Qualitative tests by X-ray fluorescence spectrometry in samples were performed to demonstrate the applicability and versatility of the project. It follows that the prototype illustrates a possible adequately to portable device for X-ray spectrometry of two-dimensional. (author)

  17. Investigation of elemental distribution in lung samples by X-ray fluorescence microtomography

    International Nuclear Information System (INIS)

    Pereira, Gabriela R.; Rocha, Henrique S.; Lopes, Ricardo T.

    2007-01-01

    X-Ray Fluorescence Microtomography (XRFCT) is a suitable technique to find elemental distributions in heterogeneous samples. While x-ray transmission microtomography provides information about the linear attenuation coefficient distribution, XRFCT allows one to map the most important elements in the sample. The x-ray fluorescence tomography is based on the use of the X-ray fluorescence emitted from the elements contained in a sample so as to give additional information to characterize the object under study. In this work a rat lung and two human lung tissue samples have been investigated in order to verify the efficiency of the system in determination of the internal distribution of detected elements in these kinds of samples and to compare the elemental distribution in the lung tissue of an old human and a fetus. The experiments were performed at the X-Ray Fluorescence beamline (XRF) of the Brazilian Synchrotron Light Source (LNLS), Campinas, Brazil. A white beam was used for the excitation of the elements and the fluorescence photons have been detected by a HPGe detector. All the tomographies have been reconstructed using a filtered-back projection algorithm. It was possible to visualize the distribution of high atomic number elements on both, artificial and tissues samples. It was compared the quantity of Zn, Cu and Fe for the lung human tissue samples and verify that these elements have a higher concentration on the fetus tissue sample than the adult tissue sample. (author)

  18. Chemical Characterization of Nuclear Materials: Development a New Combined X-Ray Fluorescence and Raman Spectrometer

    International Nuclear Information System (INIS)

    Szaloki, I.; Gerenyi, A.

    2015-01-01

    New mobile analytical device based on combination of X-ray fluorescence and Raman spectrometer has been developed for prompt and quantitative characterization of chemical component from Al to U in nuclear waste or undeclared materials. The excitation source of the X-ray fluorescence spectrometer is an air-cooled X-ray tube with Ag transmission anode. For collection of secondary X-ray photons and data processing, a compact Amptek X-ray detector system is applied with silicon drift X-ray detector. The XRF system operates in confocal mode with focal volume around 1-4 mm 3 . Varying the geometrical position and orientation of the sample optional part of its surface can be analyzed. The Raman unit includes thermoelectrically cooled laser source having 500 mW power at wavelength 785 nm. In order to obtain spectral information from sample surface a reflection-type probe is connected by optical fibres to the Raman spectrometer. A mini focusing optics is set up to the sensor-fibre that provides the system to operate as confocal optical device in reflection mode. The XRF spectrometer with X-ray detector, Raman probe and X-ray tube are mechanically fixed and hermetically connected to an aluminium chamber, which can be optionally filled with helium. The chamber is mounted on a vertical stage that provides moving it to the sample surface. A new model and computer code have been developed for XRF quantitative analysis which describes the mathematical relationship between the concentration of sample elements and their characteristic X-ray intensities. For verification of the calculations standard reference alloy samples were measured. The results was in good agreement with certified concentrations in range of 0.001-100 w%. According to these numerical results this new method is successfully applicable for quick and non-destructive quantitative analysis of waste materials without using standard samples. (author)

  19. Project Title: Radiochemical Analysis by High Sensitivity Dual-Optic Micro X-ray Fluorescence

    International Nuclear Information System (INIS)

    Havrilla, George J.; Gao, Ning

    2002-01-01

    A novel dual-optic micro X-ray fluorescence instrument will be developed to do radiochemical analysis of high-level radioactive wastes at DOE sites such as Savannah River Site and Hanford. This concept incorporates new X-ray optical elements such as monolithic polycapillaries and double bent crystals, which focus X-rays. The polycapillary optic can be used to focus X-rays emitted by the X-ray tube thereby increasing the X-ray flux on the sample over 1000 times. Polycapillaries will also be used to collect the X-rays from the excitation site and screen the radiation background from the radioactive species in the specimen. This dual-optic approach significantly reduces the background and increases the analyte signal thereby increasing the sensitivity of the analysis. A doubly bent crystal used as the focusing optic produces focused monochromatic X-ray excitation, which eliminates the bremsstrahlung background from the X-ray source. The coupling of the doubly bent crystal for monochromatic excitation with a polycapillary for signal collection can effectively eliminate the noise background and radiation background from the specimen. The integration of these X-ray optics increases the signal-to-noise and thereby increases the sensitivity of the analysis for low-level analytes. This work will address a key need for radiochemical analysis of high-level waste using a non-destructive, multi-element, and rapid method in a radiation environment. There is significant potential that this instrumentation could be capable of on-line analysis for process waste stream characterization at DOE sites

  20. Confocal total reflection X-ray fluorescence technology based on an elliptical monocapillary and a parallel polycapillary X-ray optics.

    Science.gov (United States)

    Zhu, Yu; Wang, Yabing; Sun, Tianxi; Sun, Xuepeng; Zhang, Xiaoyun; Liu, Zhiguo; Li, Yufei; Zhang, Fengshou

    2018-07-01

    A total reflection X-ray fluorescence (TXRF) spectrometer based on an elliptical monocapillary X-ray lens (MXRL) and a parallel polycapillary X-ray lens (PPXRL) was designed. This TXRF instrument has micro focal spot, low divergence and high intensity of incident X-ray beam. The diameter of the focal spot of MXRL was 16.5 µm, and the divergence of the incident X-ray beam was 3.4 mrad. We applied this TXRF instrument to the micro analysis of a single-layer film containing Ni deposited on a Si substrate by metal vapor vacuum arc ion source. Copyright © 2018 Elsevier Ltd. All rights reserved.

  1. Development of a micro-X-ray fluorescence system based on polycapillary X-ray optics for non-destructive analysis of archaeological objects

    Science.gov (United States)

    Cheng, Lin; Ding, Xunliang; Liu, Zhiguo; Pan, Qiuli; Chu, Xuelian

    2007-08-01

    A new micro-X-ray fluorescence (micro-XRF) system based on rotating anode X-ray generator and polycapillary X-ray optics has been set up in XOL Lab, BNU, China, in order to be used for analysis of archaeological objects. The polycapillary X-ray optics used here can focus the primary X-ray beam down to tens of micrometers in diameter that allows for non-destructive and local analysis of sub-mm samples with minor/trace level sensitivity. The analytical characteristics and potential of this micro-XRF system in archaeological research are discussed. Some described uses of this instrument include studying Chinese ancient porcelain.

  2. Development of a micro-X-ray fluorescence system based on polycapillary X-ray optics for non-destructive analysis of archaeological objects

    Energy Technology Data Exchange (ETDEWEB)

    Cheng Lin [Key Laboratory of Beam Technology and Material Modification of Ministry of Education, Institute of Low Energy Nuclear Physics, Beijing Normal University, Beijing Radiation Center, Beijing, 100875 (China)], E-mail: chenglin@bnu.edu.cn; Ding Xunliang; Liu Zhiguo; Pan Qiuli; Chu Xuelian [Key Laboratory of Beam Technology and Material Modification of Ministry of Education, Institute of Low Energy Nuclear Physics, Beijing Normal University, Beijing Radiation Center, Beijing, 100875 (China)

    2007-08-15

    A new micro-X-ray fluorescence (micro-XRF) system based on rotating anode X-ray generator and polycapillary X-ray optics has been set up in XOL Lab, BNU, China, in order to be used for analysis of archaeological objects. The polycapillary X-ray optics used here can focus the primary X-ray beam down to tens of micrometers in diameter that allows for non-destructive and local analysis of sub-mm samples with minor/trace level sensitivity. The analytical characteristics and potential of this micro-XRF system in archaeological research are discussed. Some described uses of this instrument include studying Chinese ancient porcelain.

  3. Surface characterization of selected polymer thin films by total-reflection x-ray fluorescence spectroscopy and x-ray reflectivity

    International Nuclear Information System (INIS)

    Innis, Vallerie Ann A.

    2006-01-01

    Development of available x-ray characterizations tools for grazing incidence techniques was done to be able to probe nano-size thin films. Alignment of a Philips x-ray powder diffractometer was improved to let it perform as an x-ray reflectometer. X-ray reflectometry was coupled with total-reflection x-ray fluorescence spectroscopy. Evaluation of the performance of this grazing incidence techniques was done by preparing polymer thin films of carboxymethylcellulose, carrageenan and polyvinylpyrrolidone (PVP). The thickness of the films were varied by varying the process parameters such as concentration, spin speed and spin time. Angle-dispersive total-reflection x-ray fluorescence spectroscopy profiles of three films showed film formation only in carrageenan and PVP. For both carrageenan and PVP, an increase in concentration yielded a corresponding increase in intensity of the fluorescent or scattered peaks. XRR profiles of carrageenan thin films yielded a mean value for the critical angle close to quartz substrate. Thickness measurements of the prepared carrageenan thin films showed that concentration was the main determinant for final film thickness over the other process parameters. Sulfur fluorescent intensity derived from the TXRF measurement showed a linear relationship with the measured thickness by XRR. For PVP, measured critical angle is lower than quartz. Poor adhesion of the polymer onto the substrate yielded a limited number of thickness measurements made from the XRR profiles. (Author)

  4. Basic design of on-line analyzer for sheet paper using X-ray fluorescence (XRF) technique

    International Nuclear Information System (INIS)

    Rony Djokorayono; Ahmad Suntoro; Ikhsan Shobari; Usep Setia Gunawan

    2016-01-01

    Basic design of on-line analyzer for sheet paper using X-ray fluorescence technique has been carried out. Compared with sampling technique, this X-ray fluorescence technique has some advantages in term of analysis accuracy and time. The design activities performed including the establishment of design requirements, functional requirements, technical requirements, technical specification, detection sub-system design, data acquisition sub-system design, and operator computer console design. This program will use silicon drift or CdTe X-ray detector to detect X-ray fluorescence emitted by elements in sheet paper due to X-ray interaction of a X-ray source, 55 Fe (Ferro-55).This basic design of on-line analyzer for sheet paper using X-ray fluorescence technique should be followed up with the development of detailed design, prototype construction, and field testing. (author)

  5. A new X-ray pinhole camera for energy dispersive X-ray fluorescence imaging with high-energy and high-spatial resolution

    Energy Technology Data Exchange (ETDEWEB)

    Romano, F.P., E-mail: romanop@lns.infn.it [IBAM, CNR, Via Biblioteca 4, 95124 Catania (Italy); INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy); Altana, C. [INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy); Dipartimento di Fisica e Astronomia, Università di Catania, Via S. Sofia 64, 95123 Catania (Italy); Cosentino, L.; Celona, L.; Gammino, S.; Mascali, D. [INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy); Pappalardo, L. [IBAM, CNR, Via Biblioteca 4, 95124 Catania (Italy); INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy); Rizzo, F. [INFN-LNS, Via S. Sofia 62, 95123 Catania (Italy); Dipartimento di Fisica e Astronomia, Università di Catania, Via S. Sofia 64, 95123 Catania (Italy)

    2013-08-01

    A new X-ray pinhole camera for the Energy Dispersive X-ray Fluorescence (ED-XRF) imaging of materials with high-energy and high-spatial resolution, was designed and developed. It consists of a back-illuminated and deep depleted CCD detector (composed of 1024 × 1024 pixels with a lateral size of 13 μm) coupled to a 70 μm laser-drilled pinhole-collimator, positioned between the sample under analysis and the CCD. The X-ray pinhole camera works in a coaxial geometry allowing a wide range of magnification values. The characteristic X-ray fluorescence is induced on the samples by irradiation with an external X-ray tube working at a maximum power of 100 W (50 kV and 2 mA operating conditions). The spectroscopic capabilities of the X-ray pinhole camera were accurately investigated. Energy response and energy calibration of the CCD detector were determined by irradiating pure target-materials emitting characteristic X-rays in the energy working-domain of the system (between 3 keV and 30 keV). Measurements were performed by using a multi-frame acquisition in single-photon counting. The characteristic X-ray spectra were obtained by an automated processing of the acquired images. The energy resolution measured at the Fe–Kα line is 157 eV. The use of the X-ray pinhole camera for the 2D resolved elemental analysis was investigated by using reference-patterns of different materials and geometries. The possibility of the elemental mapping of samples up to an area of 3 × 3 cm{sup 2} was demonstrated. Finally, the spatial resolution of the pinhole camera was measured by analyzing the profile function of a sharp-edge. The spatial resolution determined at the magnification values of 3.2 × and 0.8 × (used as testing values) is about 90 μm and 190 μm respectively. - Highlights: • We developed an X-ray pinhole camera for the 2D X-ray fluorescence imaging. • X-ray spectra are obtained by a multi-frame acquisition in single photon mode. • The energy resolution in the X-ray

  6. Determination of fission products in irradiated fuel by X-ray fluorescence

    International Nuclear Information System (INIS)

    Mogensen, M.; Als-Nielsen, J.; Hessel Andersen, N.

    1986-08-01

    X-ray fluorescence is a well established analytical tool for measuring elemental composition of fairly large (approximately 5 cm 2 ) ''cold'' samples. A version of this technique has been developed for analysis of radial distribution of fission products Xe, Cs and Ba in irradiated UO 2 fuel samples. About 0.1 mm thin slices of fuel pellets (full cross sections) are irradiated by 50 keV X-rays. The intensity of the Xe (Cs, Ba) K α fluorescence radiation generated is measured by means of a Ge detector fitted with a collimator. The slit is 0.5 mm wide in the scanning direction and 2 mm long. The measured Xe K α X-ray intensities are converted to absolute concentrations by comparing to the intensity from a Xe gas standard. In the case of Cs and Ba solid standards may be used. The X-ray fluorescence analysis is compared to other techniques used to obtain radial fission product profiles. It is shown how a combination of X-ray fluorescence and electron probe micro analysis is able to reveal the amount of Xe in the grain boundary porosities. (author)

  7. TU-G-207-03: High Spatial Resolution and High Sensitivity X-Ray Fluorescence Imaging

    International Nuclear Information System (INIS)

    Xing, L.

    2015-01-01

    Last few years has witnessed the development of novel of X-ray imaging modalities, such as spectral CT, phase contrast CT, and X-ray acoustic/fluorescence/luminescence imaging. This symposium will present the recent advances of these emerging X-ray imaging modalities and update the attendees with knowledge in various related topics, including X-ray photon-counting detectors, X-ray physics underlying the emerging applications beyond the traditional X-ray imaging, image reconstruction for the novel modalities, characterization and evaluation of the systems, and their practical implications. In addition, the concept and practical aspects of X-ray activatable targeted nanoparticles for molecular X-ray imaging will be discussed in the context of X-ray fluorescence and luminescence CT. Learning Objectives: Present background knowledge of various emerging X-ray imaging techniques, such as spectral CT, phase contrast CT and X-ray fluorescence/luminescence CT. Discuss the practical need, technical aspects and current status of the emerging X-ray imaging modalities. Describe utility and future impact of the new generation of X-ray imaging applications

  8. Choice of excitation source for determination of rare earth elements with radioisotope excited X ray fluorescence

    International Nuclear Information System (INIS)

    Zhang Quanshi; Chang Yongfu

    2000-01-01

    The comparisons of two radioisotope source ( 241 Am and 238 Pu) which are the most available in the radioisotope excited X Ray Fluorescence (XRF) analysis technique and two characteristic X ray series (KX and LX) analyzed for the determination of the rare-earth (RE) elements were investigated in detail. According to the principle of emission and detection of X ray , the relative excitation efficiencies were calculated by the some fundamental physical parameters including the photoelectric mass attenuation coefficient, the fluorescent yield, the absorption jump factor, the emission probability of the detected fluorescent line with reference to other liens of the same series etc., The advantages and disadvantages of the two conditions are discussed. These results may determine the optimal excitation and detection conditions for different rare-earth elements. The experimental results with nine rare-earth elements (Ce, Nd, Sm, Tb, Tm, Ho, Er, Yb and Lu) are in agreement with the results of theoretical calculations

  9. Analysis of eight argonne premium coal samples by X-ray fluorescence spectrometry

    Science.gov (United States)

    Evans, J.R.; Sellers, G.A.; Johnson, R.G.; Vivit, D.V.; Kent, J.

    1990-01-01

    X-ray fluorescence spectrometric methods were used in the analysis of eight Argonne Premium Coal Samples. Trace elements (Cr, Ni, Cu, Zn, Rb, Sr, Y, Zr, Nb, Ba, La, and Ce) in coal ash were determined by energy-dispersive X-ray fluorescence spectrometry; major elements (Na, Mg, Al, Si, P, S, K, Ca, Ti, Mn, and Fe) in coal ash and trace elements (Cl and P) in whole coal were determined by wavelength-dispersive X-ray fluorescence spectrometry. The results of this study will be used in a geochemical database compiled for these materials from various analytical techniques. The experimental XRF methods and procedures used to determine these major and trace elements are described.

  10. Analysis of archaeological ceramics by total-reflection X-ray fluorescence: Quantitative approaches

    International Nuclear Information System (INIS)

    Fernandez-Ruiz, R.; Garcia-Heras, M.

    2008-01-01

    This paper reports the quantitative methodologies developed for the compositional characterization of archaeological ceramics by total-reflection X-ray fluorescence at two levels. A first quantitative level which comprises an acid leaching procedure, and a second selective level, which seeks to increase the number of detectable elements by eliminating the iron present in the acid leaching procedure. Total-reflection X-ray fluorescence spectrometry has been compared, at a quantitative level, with Instrumental Neutron Activation Analysis in order to test its applicability to the study of this kind of materials. The combination of a solid chemical homogenization procedure previously reported with the quantitative methodologies here presented allows the total-reflection X-ray fluorescence to analyze 29 elements with acceptable analytical recoveries and accuracies

  11. An engineering development of fluoroscopic X-ray medical equipment based-on fluorescent screen

    International Nuclear Information System (INIS)

    Ferry Suyatno; I Putu Susila; Djoko Sukmono

    2011-01-01

    Fluoroscopic x-ray medical equipment uses fluorescent screen to capture structural image of organs. Unlike conventional x-ray equipment which uses film, in the fluoroscopic x-ray, the resulting image is visualized on the fluorescent screen and directly observed by physicians in the patients' rooms. In this study, we developed an image capture system that transforms the image on the fluorescent screen into digital data, which is then transferred to computer for visualization and further processing. By using this system, the observation of the resulting image can be done on a computer that is placed in the control room. The image can also be stored easily and at low cost compared to conventional film. The experiment shows that the system could be used to capture image of the object. However, its quality needs to be improved. In the future, the system will be modified and tested with different types of cameras to obtain better results. (author)

  12. Analysis of archaeological ceramics by total-reflection X-ray fluorescence: Quantitative approaches

    Energy Technology Data Exchange (ETDEWEB)

    Fernandez-Ruiz, R. [Servicio Interdepartamental de Investigacion, Facultad de Ciencias, Universidad Autonoma de Madrid, Modulo C-9, Laboratorio de TXRF, Crta. Colmenar, Km 15, Cantoblanco, E-28049, Madrid (Spain)], E-mail: ramon.fernandez@uam.es; Garcia-Heras, M. [Grupo de Arqueometria de Vidrios y Materiales Ceramicos, Instituto de Historia, Centro de Ciencias Humanas y Sociales, CSIC, C/ Albasanz, 26-28, 28037 Madrid (Spain)

    2008-09-15

    This paper reports the quantitative methodologies developed for the compositional characterization of archaeological ceramics by total-reflection X-ray fluorescence at two levels. A first quantitative level which comprises an acid leaching procedure, and a second selective level, which seeks to increase the number of detectable elements by eliminating the iron present in the acid leaching procedure. Total-reflection X-ray fluorescence spectrometry has been compared, at a quantitative level, with Instrumental Neutron Activation Analysis in order to test its applicability to the study of this kind of materials. The combination of a solid chemical homogenization procedure previously reported with the quantitative methodologies here presented allows the total-reflection X-ray fluorescence to analyze 29 elements with acceptable analytical recoveries and accuracies.

  13. Exploration in vivo by X-ray fluorescence (thyroid-brain)

    International Nuclear Information System (INIS)

    Delcroix, V.; Allemand, R.; Laval, M.; Dipaola, M.; Tubiana, M.

    1975-01-01

    X-ray fluorescence methods of medical exploration avoid the use of radioactive tracers and hence reduce the total dose received by the patient. In addition the collimation to the excitation source and detector respectively produces a tomographic effect which improves the spatial resolution of the system and even allows organs to be charted. The physical principles involved in X-ray fluorescence are outlined, with emphasis on the fact that the only elements useful for such applications are those of high enough atomic number to emit a fluorescence radiation of energy sufficient to pass through the tissues. The apparatus used, the excitation sources (radioactive source or X-ray tube), the detector and the measurement equipment are described. The experimental results obtained are given in two fields: measurement of blood flow in the tissues; thyroid imagery [fr

  14. Simultaneous determination of actinides by x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Cohen, V.H.

    1990-01-01

    The x-ray spectrometric simultaneous determination of uranium and plutonium in simulated Purex Process solutions is described. The method is accomplished by intensity measurements of the L α sub(1) lines. The thin film technique for sample preparation and thorium as an internal standard had been used. An evaporation technique had been also tested for low concentration uranium solutions. In the measurement range 0,05 - 130 U g/L, 0,5 - 20 Pu g/L linear calibration curves were effected. The standard deviation in the concentration range 10 to 130 g/L was 3,5%, 4% in the 1 to 10 g/L and 13% in 0,05 to 1 g/L for uranium determination and 4% for plutonium determination in the range of 1 to 20 g/L. The sensitivity of the method was about 3,62 μg to U and 3,95 μg to Pu. Uranium and plutonium do not reciprocally interfere with one another until U/Pu ≅ 90 m/m. The fission product as interfering elements were also verified. Finally, uranium and plutonium were determined in simulated Purex Process solutions within the requested accuracy for control method. (author)

  15. Determination of plutonium in nitric acid solutions using energy dispersive L X-ray fluorescence with a low power X-ray generator

    Energy Technology Data Exchange (ETDEWEB)

    Py, J. [Laboratoire Chrono-Environnement, UMR CNRS 6249, Université de Franche-Comté, 16 route de Gray, F-25030 Besançon (France); Commissariat à l’Énergie Atomique, Centre de Valduc, F-21120 Is-sur-Tille (France); Groetz, J.-E., E-mail: jegroetz@univ-fcomte.fr [Laboratoire Chrono-Environnement, UMR CNRS 6249, Université de Franche-Comté, 16 route de Gray, F-25030 Besançon (France); Hubinois, J.-C.; Cardona, D. [Commissariat à l’Énergie Atomique, Centre de Valduc, F-21120 Is-sur-Tille (France)

    2015-04-21

    This work presents the development of an in-line energy dispersive L X-ray fluorescence spectrometer set-up, with a low power X-ray generator and a secondary target, for the determination of plutonium concentration in nitric acid solutions. The intensity of the L X-rays from the internal conversion and gamma rays emitted by the daughter nuclei from plutonium is minimized and corrected, in order to eliminate the interferences with the L X-ray fluorescence spectrum. The matrix effects are then corrected by the Compton peak method. A calibration plot for plutonium solutions within the range 0.1–20 g L{sup −1} is given.

  16. Calculating the X-Ray Fluorescence from the Planet Mercury Due to High-Energy Electrons

    Science.gov (United States)

    Burbine, T. H.; Trombka, J. I.; Bergstrom, P. M., Jr.; Christon, S. P.

    2005-01-01

    The least-studied terrestrial planet is Mercury due to its proximity to the Sun, which makes telescopic observations and spacecraft encounters difficult. Our lack of knowledge about Mercury should change in the near future due to the recent launching of MESSENGER, a Mercury orbiter. Another mission (BepiColombo) is currently being planned. The x-ray spectrometer on MESSENGER (and planned for BepiColombo) can characterize the elemental composition of a planetary surface by measuring emitted fluorescent x-rays. If electrons are ejected from an atom s inner shell by interaction with energetic particles such as photons, electrons, or ions, electrons from an outer shell can transfer to the inner shell. Characteristic x-rays are then emitted with energies that are the difference between the binding energy of the ion in its excited state and that of the ion in its ground state. Because each element has a unique set of energy levels, each element emits x-rays at a unique set of energies. Electrons and ions usually do not have the needed flux at high energies to cause significant x-ray fluorescence on most planetary bodies. This is not the case for Mercury where high-energy particles were detected during the Mariner 10 flybys. Mercury has an intrinsic magnetic field that deflects the solar wind, resulting in a bow shock in the solar wind and a magnetospheric cavity. Electrons and ions accelerated in the magnetosphere tend to follow its magnetic field lines and can impact the surface on Mercury s dark side Modeling has been done to determine if x-ray fluorescence resulting from the impact of high-energy electrons accelerated in Mercury's magnetosphere can be detected by MESSENGER. Our goal is to understand how much bulk chemical information can be obtained from x-ray fluorescence measurements on the dark side of Mercury.

  17. Rapid Analysis of Copper Ore in Pre-Smelter Head Flow Slurry by Portable X-ray Fluorescence.

    Science.gov (United States)

    Burnett, Brandon J; Lawrence, Neil J; Abourahma, Jehad N; Walker, Edward B

    2016-05-01

    Copper laden ore is often concentrated using flotation. Before the head flow slurry can be smelted, it is important to know the concentration of copper and contaminants. The concentration of copper and other elements fluctuate significantly in the head flow, often requiring modification of the concentrations in the slurry prior to smelting. A rapid, real-time analytical method is needed to support on-site optimization of the smelter feedstock. A portable, handheld X-ray fluorescence spectrometer was utilized to determine the copper concentration in a head flow suspension at the slurry origin. The method requires only seconds and is reliable for copper concentrations of 2.0-25%, typically encountered in such slurries. © The Author(s) 2016.

  18. Characterization of atmospheric aerosols using Synchroton radiation total reflection X-ray fluorescence and Fe K-edge total reflection X-ray fluorescence-X-ray absorption near-edge structure

    Energy Technology Data Exchange (ETDEWEB)

    Fittschen, U.E.A. [Department of Chemistry, University of Hamburg, Martin-Luther-King-Platz 6, 20146 Hamburg (Germany)], E-mail: ursula.fittschen@chemie.uni-hamburg.de; Meirer, F. [Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Wien (Austria)], E-mail: fmeirer@ati.ac.at; Streli, C. [Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Wien (Austria)], E-mail: streli@ati.ac.at; Wobrauschek, P. [Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Wien (Austria)], E-mail: wobi@ati.ac.at; Thiele, J. [Department of Chemistry, University of Hamburg, Martin-Luther-King-Platz 6, 20146 Hamburg (Germany)], E-mail: Julian.Thiele@gmx.de; Falkenberg, G. [Hamburger Synchrotronstrahlungslabor at Deutsches Elektronen-Synchrotron DESY, Notkestr. 85, 22603 Hamburg (Germany)], E-mail: falkenbe@mail.desy.de; Pepponi, G. [ITC-irst, Via Sommarive 18, 38050 Povo (Trento) (Italy)], E-mail: pepponi@itc.it

    2008-12-15

    In this study a new procedure using Synchrotron total reflection X-ray fluorescence (SR-TXRF) to characterize elemental amounts in atmospheric aerosols down to particle sizes of 0.015 um is presented. The procedure was thoroughly evaluated regarding bounce off effects and blank values. Additionally the potential of total reflection X-ray fluorescence-X-ray absorption near edge structure (SR-TXRF-XANES) for speciation of FeII/III down to amounts of 34 pg in aerosols which were collected for 1 h is shown. The aerosols were collected in the city of Hamburg with a low pressure Berner impactor on Si carriers covered with silicone over time periods of 60 and 20 min each. The particles were collected in four and ten size fractions of 10.0-8.0 {mu}m, 8.0-2.0 {mu}m, 2.0-0.13 {mu}m 0.13-0.015 {mu}m (aerodynamic particle size) and 15-30 nm, 30-60 nm, 60-130 nm, 130-250 nm, 250-500 nm, 0.5-1 {mu}m, 1-2 {mu}m, 2-4 {mu}m, 4-8 {mu}m, 8-16 {mu}m. Prior to the sampling 'bounce off' effects on Silicone and Vaseline coated Si carriers were studied with total reflection X-ray fluorescence. According to the results silicone coated carriers were chosen for the analysis. Additionally, blank levels originating from the sampling device and the calibration procedure were studied. Blank levels of Fe corresponded to 1-10% of Fe in the aerosol samples. Blank levels stemming from the internal standard were found to be negligible. The results from the Synchroton radiation total reflection X-ray fluorescence analysis of the aerosols showed that 20 min of sampling time gave still enough sample material for elemental determination of most elements. For the determination of the oxidation state of Fe in the aerosols different Fe salts were prepared as a reference from suspensions in isopropanol. The results from the Fe K-edge Synchroton radiation total reflection X-ray fluorescence-X-ray absorption near-edge structure analysis of the aerosol samples showed that mainly Fe(III) was present in

  19. Characterization of atmospheric aerosols using Synchroton radiation total reflection X-ray fluorescence and Fe K-edge total reflection X-ray fluorescence-X-ray absorption near-edge structure

    International Nuclear Information System (INIS)

    Fittschen, U.E.A.; Meirer, F.; Streli, C.; Wobrauschek, P.; Thiele, J.; Falkenberg, G.; Pepponi, G.

    2008-01-01

    In this study a new procedure using Synchrotron total reflection X-ray fluorescence (SR-TXRF) to characterize elemental amounts in atmospheric aerosols down to particle sizes of 0.015 um is presented. The procedure was thoroughly evaluated regarding bounce off effects and blank values. Additionally the potential of total reflection X-ray fluorescence-X-ray absorption near edge structure (SR-TXRF-XANES) for speciation of FeII/III down to amounts of 34 pg in aerosols which were collected for 1 h is shown. The aerosols were collected in the city of Hamburg with a low pressure Berner impactor on Si carriers covered with silicone over time periods of 60 and 20 min each. The particles were collected in four and ten size fractions of 10.0-8.0 μm, 8.0-2.0 μm, 2.0-0.13 μm 0.13-0.015 μm (aerodynamic particle size) and 15-30 nm, 30-60 nm, 60-130 nm, 130-250 nm, 250-500 nm, 0.5-1 μm, 1-2 μm, 2-4 μm, 4-8 μm, 8-16 μm. Prior to the sampling 'bounce off' effects on Silicone and Vaseline coated Si carriers were studied with total reflection X-ray fluorescence. According to the results silicone coated carriers were chosen for the analysis. Additionally, blank levels originating from the sampling device and the calibration procedure were studied. Blank levels of Fe corresponded to 1-10% of Fe in the aerosol samples. Blank levels stemming from the internal standard were found to be negligible. The results from the Synchroton radiation total reflection X-ray fluorescence analysis of the aerosols showed that 20 min of sampling time gave still enough sample material for elemental determination of most elements. For the determination of the oxidation state of Fe in the aerosols different Fe salts were prepared as a reference from suspensions in isopropanol. The results from the Fe K-edge Synchroton radiation total reflection X-ray fluorescence-X-ray absorption near-edge structure analysis of the aerosol samples showed that mainly Fe(III) was present in all particle size fractions

  20. Abstracts of the 8th Conference on total reflection x-ray fluorescence analysis and related methods

    International Nuclear Information System (INIS)

    Wobrauschek, P.

    2000-01-01

    The 8. conference on total reflection x-ray fluorescence analysis and related methods held from 25.9 to 29.9.2000 contains 79 abstracts about x-ray fluorescence analysis (XRFA) as a powerful tool used for industrial production, geological prospecting and for environmental control. Total reflection x-ray fluorescence spectroscopy is also a tool used for chemical analysis in medicine, industry and research. (E.B.)

  1. X-Ray fluorescence determination of the mobile forms of toxic elements in meadow chernozems

    International Nuclear Information System (INIS)

    Belikov, K.N.; Blank, A.B.; Shevtsov, N.I.

    1997-01-01

    An X-ray fluorescence method for determining mobile forms of Mn, V, Cr, Co, Ni, Zn, and Pb in meadow chernozems was developed. It is based on the extraction of analytes with an acetate-ammonia buffer solution and the evaporation of the obtained filtrate with carbon powder followed by the X-ray fluorescence analysis of the dry residue. The effect of concomitants on the analytical signals of elements under determination was examined. It was suggested to spike samples with barium in order to decrease analytical errors. (author)

  2. X-ray fluorescence analysis of thulium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Khanna, P.P.

    1986-01-01

    An X-ray fluorescence spectrometric method for the analysis of thulium oxide is described. For the analysis, the sample in oxalate form is mixed with boric acid binding material and pressed into a pellet over a supporting pellet of boric acid. A wavelength dispersive Philips PW 1220 X-ray fluorescence spectrometer is used for the experiments; the minimum determination limits are 0.002per cent for Ho, Lu and Y, 0.005per cent for Dy and Er and 0.01per cent for Yb. Calculations for theoretical minimum detection limits and percent standard deviation at each concentration of the standard are carried out. (author)

  3. X-ray fluorescence analysis of lutetium oxide/oxalate for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Khanna, P.P.

    1985-01-01

    An X-ray fluorescence spectrometric method for the analysis of lutetium oxide is described. The sample in the oxalate form is mixed with boric acid binding material and pressed into a pellet over supporting pellet of boric acid. A Philips PW 1220 wavelength dispersive semiautomatic X-ray fluorescence spectrometer is used for the analysis. The minimum determination limit is 0.002 percent for Y, Er and Yb and 0.005 percent for Tm. Calculations for theoretical minimum detection limits and percent standard deviations at each concentration of the standard are carried out. (author)

  4. Improvement of graphite crystal analyzer for light elements on X-ray fluorescence holography measurement

    Science.gov (United States)

    Happo, Naohisa; Hada, Takuma; Kubota, Atsushi; Ebisu, Yoshihiro; Hosokawa, Shinya; Kimura, Koji; Tajiri, Hiroo; Matsushita, Tomohiro; Hayashi, Kouichi

    2018-05-01

    Using a graphite crystal analyzer, focused monochromatic fluorescent X-rays can be obtained on an X-ray fluorescence holography (XFH) measurement. To measure the holograms of elements lighter than Ti, we improved a cylindrical-type crystal analyzer and constructed a small C-shaped analyzer. Using the constructed C-shaped analyzer, a Ca Kα hologram of a fluorite single crystal was obtained, from which we reconstructed a clear atomic image. The XFH measurements for the K, Ca, and Sc elements become possible using the presently constructed analyzer.

  5. X-ray fluorescence analysis of bimetallic complexes on the basis of tantalocene trihydride

    International Nuclear Information System (INIS)

    Shurupova, T.I.; Sokolova, T.A.

    1989-01-01

    Methods of X-ray fluorescence determination of metals in tantalocene trihydride complexes of Cp 2 TaH 3 ·nMeHalm composition where Me=Cu, Mg, Hal=Cl, I, n=1 or 2, m=1 or 2 are developed. To obtain the form, stable in relation to the air oxygen and water vapours, the complexes were burut off up to metal oxides. Possibility of direct X-ray fluorescent determination is tested taking the most stable iodide copper-containing complex as an example

  6. Improving accuracy and capabilities of X-ray fluorescence method using intensity ratios

    Energy Technology Data Exchange (ETDEWEB)

    Garmay, Andrey V., E-mail: andrew-garmay@yandex.ru; Oskolok, Kirill V.

    2017-04-15

    An X-ray fluorescence analysis algorithm is proposed which is based on a use of ratios of X-ray fluorescence lines intensities. Such an analytical signal is more stable and leads to improved accuracy. Novel calibration equations are proposed which are suitable for analysis in a broad range of matrix compositions. To apply the algorithm to analysis of samples containing significant amount of undetectable elements a use of a dependence of a Rayleigh-to-Compton intensity ratio on a total content of these elements is suggested. The technique's validity is shown by analysis of standard steel samples, model metal oxides mixture and iron ore samples.

  7. The MicroAnalysis Toolkit: X-ray Fluorescence Image Processing Software

    International Nuclear Information System (INIS)

    Webb, S. M.

    2011-01-01

    The MicroAnalysis Toolkit is an analysis suite designed for the processing of x-ray fluorescence microprobe data. The program contains a wide variety of analysis tools, including image maps, correlation plots, simple image math, image filtering, multiple energy image fitting, semi-quantitative elemental analysis, x-ray fluorescence spectrum analysis, principle component analysis, and tomographic reconstructions. To be as widely useful as possible, data formats from many synchrotron sources can be read by the program with more formats available by request. An overview of the most common features will be presented.

  8. X-ray fluorescence analysis and optical emission spectrometry of an roman mirror from Tomis, Romania

    International Nuclear Information System (INIS)

    Belc, M.; Bogoi, M.; Ionescu, D.; Guita, D.; Caiteanu, S.; Caiteanu, D.

    2000-01-01

    The miscellaneous population of Roman Empire, their diverse cultural tradition, their ability to assimilate the roman civilization spirits, had determined a permanent reassessment superimposed upon the roman contribution. Analysis was undertaken using optical emission spectrometry and non-destructive X-ray fluorescence. X-ray fluorescence analysis is a well-established method and is often used in archaeometry and other work dealing with valuable objects pertaining to the history of art and civilization. Roman mirror analysed has been found not to be made of speculum (a high tin bronze). (authors)

  9. Analysis of selected elements in tobacco by wavelength dispersive X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Martin, J.M.

    1988-01-01

    A rapid method for the determination of 16 elements in tobacco by wavelength dispersive X-ray fluorescence spectrometry has been developed. The method is accurate and precise, and requires only 9 min per sample for quantitation. Sample preparation consists of placing a portion of dried, ground tobacco in a sample cup, and pressing at 25 tons pressure to make a compressed pellet. This pellet is then automatically analyzed by X-ray fluorescence for 16 elements. The results are stored on a computer disk for future recall and report generation. The elements are: Al, Br, Ca, Cl, Cu, Fe, K, Mg, Mn, Na, P, S, Si, Sr, Ti and Zn

  10. The measurement of X-rays radiation temperature with a new developed filter-fluorescence spectroscopy

    International Nuclear Information System (INIS)

    Zhang Chuanfei; Lin Libin; Lou Fuhong; Peng Taiping

    2001-01-01

    The author introduces how to measure the energy spectra of X-rays by filter-fluorescence spectroscopy. The design principle and structure of new-developed double diaphragms and filter-fluorescence spectroscopy with 5 channels are depicted. The parameters of optimized spectroscopy by numerical method are given. The filter-fluorescence spectroscopy designed according as Rousseau balance principle improves signal-noises ratio

  11. X-ray fluorescence in IAEA Member States: Peru[X-ray fluorescence in the IAEA and its Member States

    Energy Technology Data Exchange (ETDEWEB)

    Olivera, P [Instituto Peruano de Energia Nuclear, Lab. de Fluorescencia de Rayos X, Lima (Peru)

    2005-12-15

    The application of the XRF technique performed in our laboratories in the field of archaeology began in 1994. It started with the analysis of different samples, as an incisive tooth, a fragment of occipital, a rib fragment, a segment of dorsal vertebra and small glasses found in the internal surface of a Nazca Culture mummy. In the paleopathological study performed by Dr. Guido Lombardi a {sup 109}Cd excitation source and an X ray spectrometry system were used. The elemental concentrations of Ca, Fe, Zn, and Sr were determined for the paleonutricion studies by using the relationships of Zn/Ca and Sr/Ca. The XRF technique also contributed to the mycobacterium tuberculosis detection in a mummy, supplementing other non destructive tests carried out previously. In recent years, 29 right clavicles belonging to mature individuals were analyzed out of 143 found in the Villa El Salvador area. Sr and Zn concentrations were used to determine the relative proportion of vegetable and animal foods in the population's diet under study. It was concluded that the old residents of this town manifested an omnivorous feeding with carnivorous tendency, due to consumption of products of marine origin. Ceramics have also been analyzed to determine the chemical composition of the paste which was used in the production process. 39 fragments of ceramic from the place called Lomo de Corvina of Villa El Salvador, low valley in Lurin town were analyzed. Thalysis of these samples was focused on quantitative determination of Ti, Fe, Rb, Sr, Y, Zr and Nb, to complement the results obtained by neutron activation analysis. Identification of the pigments used for production of art objects is of great importance for its characterization, authentication and/or restoration. For this reason, we are currently working on identification of pigments used in the decoration of archaeological ceramics. This work started from construction of a portable XRF spectrometer based on a small size, low power X ray

  12. Application of x-ray fluorescence to the measurement of additives in paper

    International Nuclear Information System (INIS)

    Buchnea, A.; McNelles, L.A.; Sinclair, A.H.; Hewitt, J.S.

    1976-01-01

    Titanium dioxide content in paper was measured by x-ray fluorescence analysis using an 55 Fe source and an x-ray proportional counter to determine the feasibility of an on-line instrument. X-ray calibration curves for 60- and 100-g/m 2 paper samples were obtained using neutron activation to measure the titanium dioxide concentration. The predictions of a simple model were in good agreement with the experimental calibration curves. The measurements and calculations were extended to investigate the effects of clay and moisture. The presence of clay has a significant effect on the x-ray fluorescence determination of the titanium dioxide concentration; however, this can be well accounted for by the model. The calculations indicated that the effect of typical moisture levels on the titanium dioxide determination was small and can be ignored. It is not possible to measure the clay content by x-ray fluorescence; however, preliminary results for the determination of calcium carbonate concentration are promising

  13. Analysis of kiwi fruit (Accented deliciosa) by energy dispersive X-ray fluorescence spectra

    International Nuclear Information System (INIS)

    Oliveira, Ana Claudia S.; Oliveira, Marcia L. de; Silva, Lucia C.A.S.; Arthur, Valter; Almeida, Eduardo

    2011-01-01

    The search for a healthy life has led consumers to eat fruits and vegetables in place of manufactured products, however, the demand for minimally processed products has evolved rapidly. The kiwi has at least eight nutrients beneficial to health: calcium, magnesium, manganese, phosphorus, iron, potassium, sodium and has also high vitamin C, which has wide acceptance in consumer markets. Energy dispersive spectroscopy X-ray (EDX) is the analytical technique used for elemental analysis or chemical characterization of a sample. It is a variant of fluorescence spectroscopy X-ray based on the sample through an investigation of interactions between electromagnetic radiation and matter, analyzing X-rays emitted by matter in response to being struck by charged particles. The aim of this study were to determine potassium, calcium, iron and bromine (K, Ca, Fe and Br, respectively) present in kiwifruit using the technique of fluorescence X-ray energy dispersive (EDXRF). Kiwifruit were peeled, washed and cut into slices and freeze-dried. After drying the sample was held digestion and subsequent reading of the same equipment in the X-ray fluorescence energy dispersive (EDXRF). The results indicated that the contents of potassium, calcium, iron and bromine are present in kiwifruit as expected when compared to Brazilian Table of Food Composition. (author)

  14. Three dimensional subsurface elemental identification of minerals using confocal micro-X-ray fluorescence and micro-X-ray computed tomography

    International Nuclear Information System (INIS)

    Cordes, Nikolaus L.; Seshadri, Srivatsan; Havrilla, George J.; Yuan, Xiaoli; Feser, Michael; Patterson, Brian M.

    2015-01-01

    Current non-destructive elemental characterization methods, such as scanning electron microscopy-based energy dispersive spectroscopy (SEM–EDS) and micro-X-ray fluorescence spectroscopy (MXRF), are limited to either elemental identification at the surface (SEM–EDS) or suffer from an inability to discriminate between surface or depth information (MXRF). Thus, a non-destructive elemental characterization of individual embedded particles beneath the surface is impossible with either of these techniques. This limitation can be overcome by using laboratory-based 3D confocal micro-X-ray fluorescence spectroscopy (confocal MXRF). This technique utilizes focusing optics on the X-ray source and detector which allows for spatial discrimination in all three dimensions. However, the voxel-by-voxel serial acquisition of a 3D elemental scan can be very time-intensive (~ 1 to 4 weeks) if it is necessary to locate individual embedded particles of interest. As an example, if each point takes a 5 s measurement time, a small volume of 50 × 50 × 50 pixels leads to an acquisition time of approximately 174 h, not including sample stage movement time. Initially screening the samples for particles of interest using micro-X-ray computed tomography (micro-CT) can significantly reduce the time required to spatially locate these particles. Once located, these individual particles can be elementally characterized with confocal MXRF. Herein, we report the elemental identification of high atomic number surface and subsurface particles embedded in a mineralogical matrix by coupling micro-CT and confocal MXRF. Synergistically, these two X-ray based techniques first rapidly locate and then elementally identify individual subsurface particles. - Highlights: • Coupling of confocal X-ray fluorescence spectroscopy and X-ray computed tomography • Qualitative elemental identification of surface and subsurface mineral particles • Non-destructive particle size measurements • Utilization of

  15. Three dimensional subsurface elemental identification of minerals using confocal micro-X-ray fluorescence and micro-X-ray computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Cordes, Nikolaus L., E-mail: ncordes@lanl.gov [Polymers and Coatings Group, Material Science and Technology Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States); Seshadri, Srivatsan, E-mail: srivatsan.seshadri@zeiss.com [Carl Zeiss X-ray Microscopy, Inc., Pleasanton, CA 94588 (United States); Havrilla, George J. [Chemical Diagnostics and Engineering, Chemistry Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States); Yuan, Xiaoli [Julius Kruttschnitt Mineral Research Centre, University of Queensland, Indooroopilly, Brisbane, QLD 4068 (Australia); Feser, Michael [Carl Zeiss X-ray Microscopy, Inc., Pleasanton, CA 94588 (United States); Patterson, Brian M. [Polymers and Coatings Group, Material Science and Technology Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States)

    2015-01-01

    Current non-destructive elemental characterization methods, such as scanning electron microscopy-based energy dispersive spectroscopy (SEM–EDS) and micro-X-ray fluorescence spectroscopy (MXRF), are limited to either elemental identification at the surface (SEM–EDS) or suffer from an inability to discriminate between surface or depth information (MXRF). Thus, a non-destructive elemental characterization of individual embedded particles beneath the surface is impossible with either of these techniques. This limitation can be overcome by using laboratory-based 3D confocal micro-X-ray fluorescence spectroscopy (confocal MXRF). This technique utilizes focusing optics on the X-ray source and detector which allows for spatial discrimination in all three dimensions. However, the voxel-by-voxel serial acquisition of a 3D elemental scan can be very time-intensive (~ 1 to 4 weeks) if it is necessary to locate individual embedded particles of interest. As an example, if each point takes a 5 s measurement time, a small volume of 50 × 50 × 50 pixels leads to an acquisition time of approximately 174 h, not including sample stage movement time. Initially screening the samples for particles of interest using micro-X-ray computed tomography (micro-CT) can significantly reduce the time required to spatially locate these particles. Once located, these individual particles can be elementally characterized with confocal MXRF. Herein, we report the elemental identification of high atomic number surface and subsurface particles embedded in a mineralogical matrix by coupling micro-CT and confocal MXRF. Synergistically, these two X-ray based techniques first rapidly locate and then elementally identify individual subsurface particles. - Highlights: • Coupling of confocal X-ray fluorescence spectroscopy and X-ray computed tomography • Qualitative elemental identification of surface and subsurface mineral particles • Non-destructive particle size measurements • Utilization of

  16. The application of a microstrip gas counter to energy-dispersive x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Veloso, J.F.C.A.; Santos, J.M.F. dos; Conde, C.A.N.

    1996-01-01

    Performance characteristics of a microstrip gas counter operated as a x-ray fluorescence spectrometer are reported. Gas amplification as a function of microstrip anode-cathode voltage was measured, and the breakdown threshold voltage was determined in pure xenon. The detector temporal stability and the effect of gas purity were assessed. Energy resolution and linearity, detection efficiency, and uniformity of spatial response in the 2- to 60-keV x-ray energy range were determined from the pulse-height distributions of the fluorescence x-ray spectra induced in a variety of single- and multi-element sample materials. Energy resolution similar to conventional proportional counters was achieved at 6 keV

  17. Total-reflection X-ray fluorescence analysis of Austrian wine

    Energy Technology Data Exchange (ETDEWEB)

    Gruber, X. [Atominstitut der Osterreichischen Universitaeten, 1020 Vienna (Austria); Kregsamer, P. [Atominstitut der Osterreichischen Universitaeten, 1020 Vienna (Austria); Wobrauschek, P. [Atominstitut der Osterreichischen Universitaeten, 1020 Vienna (Austria); Streli, C. [Atominstitut der Osterreichischen Universitaeten, 1020 Vienna (Austria)]. E-mail: streli@ati.ac.at

    2006-11-15

    The concentration of major, minor and trace elements in Austrian wine was determined by total-reflection X-ray fluorescence using gallium as internal standard. A multi-elemental analysis was possible by pipetting 6 {mu}l of wine directly on the reflector and drying. Total-reflection X-ray fluorescence analysis was performed with Atomika EXTRA II A (Cameca) X-rays from a Mo tube with a high-energy cut-off at 20 keV in total-reflection geometry. The results showed that it was possible to identify only by the elemental analysis as fingerprint the vineyards and year of vintage among 11 different wines.

  18. Total-reflection X-ray fluorescence analysis of Austrian wine

    International Nuclear Information System (INIS)

    Gruber, X.; Kregsamer, P.; Wobrauschek, P.; Streli, C.

    2006-01-01

    The concentration of major, minor and trace elements in Austrian wine was determined by total-reflection X-ray fluorescence using gallium as internal standard. A multi-elemental analysis was possible by pipetting 6 μl of wine directly on the reflector and drying. Total-reflection X-ray fluorescence analysis was performed with Atomika EXTRA II A (Cameca) X-rays from a Mo tube with a high-energy cut-off at 20 keV in total-reflection geometry. The results showed that it was possible to identify only by the elemental analysis as fingerprint the vineyards and year of vintage among 11 different wines

  19. Characterization of LiF-based soft X-ray imaging detectors by confocal fluorescence microscopy

    International Nuclear Information System (INIS)

    Bonfigli, F; Gaudio, P; Lupelli, I; Nichelatti, E; Richetta, M; Vincenti, M A; Montereali, R M

    2010-01-01

    X-ray microscopy represents a powerful tool to obtain images of samples with very high spatial resolution. The main limitation of this technique is represented by the poor spatial resolution of standard imaging detectors. We proposed an innovative high-performance X-ray imaging detector based on the visible photoluminescence of colour centres in lithium fluoride. In this work, a confocal microscope in fluorescence mode was used to characterize LiF-based imaging detectors measuring CC integrated visible fluorescence signals of LiF crystals and films (grown on several kinds of substrates) irradiated by soft X-rays produced by a laser plasma source in different exposure conditions. The results are compared with the CC photoluminescence spectra measured on the same samples and discussed.

  20. Characterization of Roman glass tesserae from the Coriglia excavation site (Italy) via energy-dispersive X-ray fluorescence spectrometry and Raman spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Donais, Mary Kate; Sparks, Andrew; Redente, Monica [Saint Anselm College, Department of Chemistry, Manchester, NH (United States); Pevenage, Jolien van; Moens, Luc; Vincze, Laszlo [Ghent University, Department of Analytical Chemistry, Ghent (Belgium); George, David B. [Saint Anselm College, Department of Classics, Manchester, NH (United States); Vandenabeele, Peter [Ghent University, Department of Archaeology, Ghent (Belgium)

    2016-12-15

    The combined use of handheld energy-dispersive X-ray fluorescence spectrometry, Raman spectroscopy, and micro-energy-dispersive X-ray fluorescence spectrometry permitted the characterization of Roman glass tesserae excavation from the Coriglia (Italy) archeological site. Analyses of ten different glass colors were conducted as spot analyses on intact samples and as both spot analyses and line scans on select cross-sectioned samples. The elemental and molecular information gained from these spectral measurements allowed for the qualitative chemical characterization of the bulk glass, decolorants, opacifiers, and coloring agents. The use of an antimony opacifier in many of the samples supports the late Imperial phasing as determined through numismatic, fresco, ceramics, and architectural evidence. And dealinization of the exterior glass layers caused by the burial environment was confirmed. (orig.)

  1. Characterization of contaminated soils using a x-ray fluorescence equipment

    International Nuclear Information System (INIS)

    Quindos, L. S.; Fuente, I.; Sainz, C.; Queral, I.

    2006-01-01

    Fast and accurate material identification is required in many areas of the metal industry as well as surveillance of contaminated areas. The portable X-ray analyzer described in this paper consists of a hand-held probe and electronic unit. The system is configured with three excitation sources, such as ''241 Am, ''109Cd and ''55 Fe, which provides excellent performance for common elements to be controlled in contaminated soils. Data about measurements developed with this device in field as well as in the laboratory are shown in this paper. (Author)

  2. Proton induced X-Ray fluorescence study as a tool trace element analysis

    International Nuclear Information System (INIS)

    El-Kady, Ahmed A.

    1978-01-01

    Usefulness and limitations of trace elemental analysis by high energy charged particles and photon induced X-ray have been discussed. Comparison with the well established neutron activation analysis technique is also given. Back-ground radiation due to bremsstrahlung from secondary electrons and due to charged particle bremsstrahlung have been reviewed for different projectiles. The sensitivity of elemental analysis by proton induced X-ray fluorescence have been examined by measuring the characteristic X-ray emission cross section for K and L transitions of many elements and for different proton energies and compared with theroretical values. The discussion given in this report show that with suitable proton generator and a high resolution X-ray detector, proton X-ray fluorescence technique is capable of analyzing many elements simultaneously at the part per million level and offers a rapid and reliable method for trace element analysis. Data on water, blood and tissue samples given in this report are few examples of many possible applications

  3. Development of an X-ray fluorescence holographic measurement system for protein crystals

    Energy Technology Data Exchange (ETDEWEB)

    Sato-Tomita, Ayana, E-mail: ayana.sato@jichi.ac.jp, E-mail: shibayam@jichi.ac.jp, E-mail: hayashi.koichi@nitech.ac.jp; Shibayama, Naoya, E-mail: ayana.sato@jichi.ac.jp, E-mail: shibayam@jichi.ac.jp, E-mail: hayashi.koichi@nitech.ac.jp; Okabe, Takahiro [Division of Biophysics, Department of Physiology, Jichi Medical University, Yakushiji, Shimotsuke 329-0498 (Japan); Happo, Naohisa [Department of Computer and Network Engineering, Graduate School of Information Sciences, Hiroshima City University, Asa-Minami-Ku, Hiroshima 731-3194 (Japan); Kimura, Koji [Department of Physical Science and Engineering, Nagoya Institute of Technology, Gokiso, Showa, Nagoya 466-8555 (Japan); Matsushita, Tomohiro [Japan Synchrotron Radiation Research Institute (JASRI), SPring-8, Sayo, Hyogo 679-5198 (Japan); Park, Sam-Yong [Drug Design Laboratory, Department of Medical Life Science, Yokohama City University, Suehiro, Tsurumi, Yokohama 230-0045 (Japan); Sasaki, Yuji C. [Department of Advanced Material Science, Graduate School of Frontier Science, The University of Tokyo, Kashiwanoha, Kashiwa 277-8561 (Japan); Hayashi, Kouichi, E-mail: ayana.sato@jichi.ac.jp, E-mail: shibayam@jichi.ac.jp, E-mail: hayashi.koichi@nitech.ac.jp [Department of Physical Science and Engineering, Nagoya Institute of Technology, Gokiso, Showa, Nagoya 466-8555 (Japan); Frontier Research Institute for Materials Science, Nagoya Institute of Technology, Gokiso, Showa, Nagoya 466-8555 (Japan)

    2016-06-15

    Experimental procedure and setup for obtaining X-ray fluorescence hologram of crystalline metalloprotein samples are described. Human hemoglobin, an α{sub 2}β{sub 2} tetrameric metalloprotein containing the Fe(II) heme active-site in each chain, was chosen for this study because of its wealth of crystallographic data. A cold gas flow system was introduced to reduce X-ray radiation damage of protein crystals that are usually fragile and susceptible to damage. A χ-stage was installed to rotate the sample while avoiding intersection between the X-ray beam and the sample loop or holder, which is needed for supporting fragile protein crystals. Huge hemoglobin crystals (with a maximum size of 8 × 6 × 3 mm{sup 3}) were prepared and used to keep the footprint of the incident X-ray beam smaller than the sample size during the entire course of the measurement with the incident angle of 0°-70°. Under these experimental and data acquisition conditions, we achieved the first observation of the X-ray fluorescence hologram pattern from the protein crystals with minimal radiation damage, opening up a new and potential method for investigating the stereochemistry of the metal active-sites in biomacromolecules.

  4. Development of an X-ray fluorescence holographic measurement system for protein crystals

    International Nuclear Information System (INIS)

    Sato-Tomita, Ayana; Shibayama, Naoya; Okabe, Takahiro; Happo, Naohisa; Kimura, Koji; Matsushita, Tomohiro; Park, Sam-Yong; Sasaki, Yuji C.; Hayashi, Kouichi

    2016-01-01

    Experimental procedure and setup for obtaining X-ray fluorescence hologram of crystalline metalloprotein samples are described. Human hemoglobin, an α_2β_2 tetrameric metalloprotein containing the Fe(II) heme active-site in each chain, was chosen for this study because of its wealth of crystallographic data. A cold gas flow system was introduced to reduce X-ray radiation damage of protein crystals that are usually fragile and susceptible to damage. A χ-stage was installed to rotate the sample while avoiding intersection between the X-ray beam and the sample loop or holder, which is needed for supporting fragile protein crystals. Huge hemoglobin crystals (with a maximum size of 8 × 6 × 3 mm"3) were prepared and used to keep the footprint of the incident X-ray beam smaller than the sample size during the entire course of the measurement with the incident angle of 0°-70°. Under these experimental and data acquisition conditions, we achieved the first observation of the X-ray fluorescence hologram pattern from the protein crystals with minimal radiation damage, opening up a new and potential method for investigating the stereochemistry of the metal active-sites in biomacromolecules.

  5. Chemical analysis by X-ray fluorescence, of niobium in high-strength plate steels

    International Nuclear Information System (INIS)

    Iozzi, F.B.; Dias, M.J.P.

    1981-01-01

    The use of X-ray fluorescence spectrometry in quantitative analysis of niobium in steels, as an alternative solution for optical emission spectrometry, in the rapid chemical control of steel fabrication by LD type converters, is presented. (M.C.K.) [pt

  6. Precision evaluation of pressed pastille preparation different methods for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Lima, Raquel Franco de Souza; Melo Junior, Germano; Sa, Jaziel Martins

    1997-01-01

    This work relates the comparison between the results obtained with the two different methods of preparing pressed pastilles from the crushed sample. In this study, the reproductivity is evaluated, aiming to define the method that furnishes a better analytic precision. These analyses were realized with a X-ray fluorescence spectrometer at the Geology Department of the Federal University of Rio Grande do Norte

  7. Analysis of River sediments from the Tigre river (Venezuela) by radioisotope excited x-ray fluorescence

    International Nuclear Information System (INIS)

    LaBrecque, J.J.; Rosales, P.A.; Schorin, H.

    1985-01-01

    This paper describes qualitative elemental scans by both energy dispersive (radioisotope excited) and conventional wavelength dispersive x-ray fluorescence of different grain size fractions of river sediments. An internal standard thin-film technique was used. The precision of Rb, Sr, Y and Zr determination for SY-3 standard reference rock and one real sample for five independently prepared samples is demonstrated

  8. Rapid extraction and x-ray fluorescence determination of gold in rock

    International Nuclear Information System (INIS)

    Lobanov, F.I.; Logunova, S.A.; Popov, A.M.; Krasnopevtseva, E.V.

    1994-01-01

    The optimal conditions for gold extractional recovery by melts of aliphatic monocarboxylic acids mixtures (C ≥ 16) with additions of di-2-ethylhexyldithiophosphoric acid, di-n-oxtylsulfide and alkylaniline were determined. A rapid method of extraction and X-ray fluorescence detection of gold in the presence of considerable amounts of Fe, Cu, Pb, Cd, Zn and Bi was developed

  9. Microanalysis of old violin varnishes by total-reflection X-ray fluorescence

    Science.gov (United States)

    von Bohlen, Alex; Meyer, Friedrich

    1997-07-01

    Total reflection X-ray fluorescence was used to characterize elements (with Z>13) contained in varnishes applied by prominent violin makers during the last five centuries. Direct analyses of small flakes with masses varnish. Higher amounts of Fe, As and Pb were found in old products, Mn, Co, Cu, Zn and Pb were used in more recent varnishes.

  10. X-ray fluorescence analysis of ancient and medieval brass artifacts from south Moravia

    Czech Academy of Sciences Publication Activity Database

    Hložek, M.; Komoróczy, Balázs; Trojek, T.

    2012-01-01

    Roč. 7, č. 70 (2012), s. 1250-1253 ISSN 0969-8043 Institutional research plan: CEZ:AV0Z80010507 Institutional support: RVO:68081758 Keywords : x-ray fluorescence analysis * brass * Moravia Subject RIV: AC - Archeology, Anthropology, Ethnology Impact factor: 1.179, year: 2012

  11. Confirmation of molecular formulas of metallic complexes through X-ray fluorescence quantitative analysis

    International Nuclear Information System (INIS)

    Filgueiras, C.A.L.; Marques, E.V.; Machado, R.M.

    1984-01-01

    X-ray fluorescence spectrophotometry was employed to determined the metal content in a series of five transition element complexes (Mn, Ti, Zn, V). The results confirmed the molecular formulas of these complexes, already proposed on the basis of elemental microanalysis, solution condutimetry and other analytical methods. (C.L.B.) [pt

  12. X-ray fluorescence determination of cobalt in iron-manganese oceanic concretions

    International Nuclear Information System (INIS)

    Ivanenko, V.V.; Kustov, V.N.; Metelev, A.Yu.; Rakita, K.A.

    1989-01-01

    A method was developed for resolution of weak analytical lines for elements determined by radionuclide-excited X-ray fluorescence multi-element analysis. The method was used aboart for determining cobalt and some other commercially valuable elements in iron-manganese concretions of Pacific ocean 109 Cd was used as an ionizing radiation source

  13. Some applications of x-ray fluorescence spectrography to the determination of uranium and thorium

    International Nuclear Information System (INIS)

    Jones, R.W.

    1959-04-01

    Several methods for the determination of uranium and thorium by X-ray fluorescence spectrography are described. In pure solutions the sensitivity for these elements is 5-10 ppm. For solutions containing gross concentrations of impurities, strontium is added as an internal standard. Precision and accuracy of the determinations are about 1% when working in the optimum concentration range. (author)

  14. Comparative analysis of toxic elements in snuff by analytical techniques of X-ray fluorescence

    International Nuclear Information System (INIS)

    Mendoza, Mario; Olivera, Paula

    2013-01-01

    Six samples of different commercial brands of cigarettes expended in the Peruvian market have been analyzed along with two IAEA certified reference material using the technique of X-ray fluorescence energy dispersive. The results obtained in the study showed the presence of toxic metals such as Cr, Mn, Fe, Cu and Zn.

  15. The X-ray fluorescent method for determination of total sulphur in bituminous coals

    International Nuclear Information System (INIS)

    Widowska-Kusmierska, J.; Siess, K.

    1979-01-01

    The X-ray fluorescent technique for the determination of total sulphur covering concentrations from 0,1 to 10% has been applied for bituminous coals showing a great variability in qualitative and quantitative composition of mineral matter (ash). The described method is a quick one giving results during one hour. The obtained good accuracy of determinations gives prospects for wide industrial application. (author)

  16. Energy-Dispersive X-Ray Fluorescence Spectrometry: A Long Overdue Addition to the Chemistry Curriculum

    Science.gov (United States)

    Palmer, Peter T.

    2011-01-01

    Portable Energy-Dispersive X-Ray Fluorescence (XRF) analyzers have undergone significant improvements over the past decade. Salient advantages of XRF for elemental analysis include minimal sample preparation, multielement analysis capabilities, detection limits in the low parts per million (ppm) range, and analysis times on the order of 1 min.…

  17. Photon induced x-ray fluorescence analysis using energy dispersive detector and dichotomous sampler

    International Nuclear Information System (INIS)

    Jaklevic, J.M.; Loo, B.W.; Goulding, F.S.

    1976-01-01

    Operating experience in using the photon-excited energy-dispersive x-ray fluorescence analysis system has demonstrated the applicability of this technique to large-scale air-sampling networks. This experience has shown that it is possible to perform automatic sampling and analysis of aerosol particulates at a sensitivity and accuracy more than adequate for most air pollution studies

  18. Fast, versatile x-ray fluorescence method for measuring tin in impregnated wood

    DEFF Research Database (Denmark)

    Drabæk, I.; Christensen, Leif Højslet

    1985-01-01

    The present paper describes an energy-dispersive x-ray fluorescence method for measuring tin in bis(tri-n-butyl)tin-oxide impregnated wood. The proposed method is of the backscatter/fundamental parameter type. Its versatility, precision, and accuracy is demonstrated by analyses of eleven samples...

  19. Total reflection X-ray fluorescence with synchrotron radiation applied to biological and environmental samples

    International Nuclear Information System (INIS)

    Simabuco, S.M.; Matsumoto, E.; Jesus, E.F.O.; Lopes, R.T.; Perez, C.; Nascimento Filho, V.F.; Costa, R.S.S.; Tavares do Carmo, M.G.; Saunders, C.

    2001-01-01

    Full text: The Total Reflection X-ray Fluorescence has been applied for trace elements in water and aqueous solutions, environmental samples and biological materials after sample preparation and to surface analysis of silicon wafers. The present paper shows some results of applications for rainwater, atmospheric particulate material, colostrum and nuclear samples. (author)

  20. X-ray fluorescence analysis study. Final report, December 1, 1970-December 31, 1977

    Energy Technology Data Exchange (ETDEWEB)

    Kneip, T J; Laurer, G R

    1978-01-01

    This report has described the most significant experiments and the results obtained, during the development of a system for the detection and measurement of Pb in blood using radioisotope-excited x-ray fluorescence analysis, over the contract period. Briefly, the report described: detector selection; source selection; source-sample-detector geometry; sample preparation; system calibration; and separation technique. (PSB)

  1. X-ray fluorescence in some rare earth and high Z elements excited ...

    Indian Academy of Sciences (India)

    section and X-ray fluorescence yield values for different elements at various photoion- ization energies ... In the present method, the K-shell vacancies are created by photons and filled by outer electrons leading to ... The well type detector pro-.

  2. Determination of rhenium in molybdenite by X-ray fluorescence: A combined chemical-spectrometric technique.

    Science.gov (United States)

    Solt, M W; Wahlberg, J S; Myers, A T

    1969-01-01

    Rhenium in molybdenite is separated from molybdenum by distillation of rhenium heptoxide from a perchloric-sulphuric acid mixture. It is concentrated by precipitation of the sulphide and then determined by X-ray fluorescence. From 3 to 1000 microg of rhenium can be measured with a precision generally within 2%. The procedure tolerates larger amounts of molybdenum than the usual colorimetric methods.

  3. Determination of rhenium in molybdenite by X-ray fluorescence. A combined chemical-spectrometric technique

    Science.gov (United States)

    Solt, M.W.; Wahlberg, J.S.; Myers, A.T.

    1969-01-01

    Rhenium in molybdenite is separated from molybdenum by distillation of rhenium heptoxide from a perchloric-sulphuric acid mixture. It is concentrated by precipitation of the sulphide and then determined by X-ray fluorescence. From 3 to 1000 ??g of rhenium can be measured with a precision generally within 2%. The procedure tolerates larger amounts of molybdenum than the usual colorimetric methods. ?? 1969.

  4. X-ray fluorescence activities at Saha Institute of Nuclear Physics, India

    Indian Academy of Sciences (India)

    X-ray fluorescence (EDXRF) system was developed to analyse alloys by exposing them directly under the tube bremsstrahlung .... A detailed description of the analysis of these data is given in [11]. ..... samples were first dried, ground in an agate mortar with pestle, and if necessary, mixed with a binder to make pellets in a ...

  5. Some aspects of detectors and electronics for x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Goulding, F.S.

    1976-08-01

    Some of the less recognized and potentially important parameters of the electronics and detectors used in X-ray fluorescence spectrometers are discussed. Detector factors include window (dead-layer) effects, time-dependent background and excess background. Noise parameters of field-effect transistors and time-variant pulse shaping are also discussed

  6. X-ray fluorescence camera for imaging of iodine media in vivo.

    Science.gov (United States)

    Matsukiyo, Hiroshi; Watanabe, Manabu; Sato, Eiichi; Osawa, Akihiro; Enomoto, Toshiyuki; Nagao, Jiro; Abderyim, Purkhet; Aizawa, Katsuo; Tanaka, Etsuro; Mori, Hidezo; Kawai, Toshiaki; Ehara, Shigeru; Sato, Shigehiro; Ogawa, Akira; Onagawa, Jun

    2009-01-01

    X-ray fluorescence (XRF) analysis is useful for measuring density distributions of contrast media in vivo. An XRF camera was developed for carrying out mapping for iodine-based contrast media used in medical angiography. Objects are exposed by an X-ray beam from a cerium target. Cerium K-series X-rays are absorbed effectively by iodine media in objects, and iodine fluorescence is produced from the objects. Next, iodine Kalpha fluorescence is selected out by use of a 58-microm-thick stannum filter and is detected by a cadmium telluride (CdTe) detector. The Kalpha rays are discriminated out by a multichannel analyzer, and the number of photons is counted by a counter card. The objects are moved and scanned by an x-y stage in conjunction with a two-stage controller, and X-ray images obtained by iodine mapping are shown on a personal computer monitor. The scan pitch of the x and y axes was 2.5 mm, and the photon counting time per mapping point was 2.0 s. We carried out iodine mapping of non-living animals (phantoms), and iodine Kalpha fluorescence was produced from weakly remaining iodine elements in a rabbit skin cancer.

  7. Modelling of a total reflection X-ray fluorescence (TXRF) system ...

    African Journals Online (AJOL)

    The simulation of the different stages involved in x-ray fluorescence emissions was carried out by writing a suite of computer programs using FORTRAN programming language. These computer simulated XRF stages were then integrated together to generate a general robust model which was run with the digital visual ...

  8. Determination of Fe and Zn in healing plants by radionuclide X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Harangozo, M.; Toelgyessy, J.; Tomecek, O.; Ruzicka, I.; Cejpek, K.

    1999-01-01

    Radionuclide X-ray fluorescence method was used for the determination of Fe and Zn in healing plants (Sage, Peppermint, Stinging, Common Agrimony, Milfoil, Ribwort, Tansy, White Dead-Nettle). 238 Pu exciting source and Si/Li semiconductor detector were used for the determination. (author)

  9. X-ray fluorescence in some medium-Z elements excited by 59.5 keV photons

    International Nuclear Information System (INIS)

    Han, I.; Shahin, M.; Demir, L.; Narmanli, E.

    2010-01-01

    K X-ray fluorescence parameters cross sections and average shell fluorescence yields) for selected ten elements in the atomic range 42 ≤ Z ≤ 66 have been experimentally determined at photon excitation energy of 59.5 keV. K X-rays emitted from the samples have been counted by a Si (Li) detector. The K spectra for investigated elements have been derived from the measured K shell X-ray spectra by peak fitting process. Experimental results of K X-ray fluorescence parameters have been compared with theory. In general there is an agreement within the standard uncertainties of the experimental and theoretical values

  10. MCNP calculation for calibration curve of X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Tan Chunming; Wu Zhifang; Guo Xiaojing; Xing Guilai; Wang Zhentao

    2011-01-01

    Due to the compositional variation of the sample, linear relationship between the element concentration and fluorescent intensity will not be well maintained in most X-ray fluorescence analysis. To overcome this, we use MCNP program to simulate fluorescent intensity of Fe (0∼100% concentration range) within binary mixture of Cr and O which represent typical strong absorption and weak absorption conditions respectively. The theoretic calculation shows that the relationship can be described as a curve determined by parameter p and value of p can be obtained with given absorption coefficient of substrate elements and element under detection. MCNP simulation results are consistent with theoretic calculation. Our research reveals that MCNP program can calculate the Calibration Curve of X-ray fluorescence very well. (authors)

  11. Improving x-ray fluorescence signal for benchtop polychromatic cone-beam x-ray fluorescence computed tomography by incident x-ray spectrum optimization: a Monte Carlo study.

    Science.gov (United States)

    Manohar, Nivedh; Jones, Bernard L; Cho, Sang Hyun

    2014-10-01

    To develop an accurate and comprehensive Monte Carlo (MC) model of an experimental benchtop polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) setup and apply this MC model to optimize incident x-ray spectrum for improving production/detection of x-ray fluorescence photons from gold nanoparticles (GNPs). A detailed MC model, based on an experimental XFCT system, was created using the Monte Carlo N-Particle (MCNP) transport code. The model was validated by comparing MC results including x-ray fluorescence (XRF) and scatter photon spectra with measured data obtained under identical conditions using 105 kVp cone-beam x-rays filtered by either 1 mm of lead (Pb) or 0.9 mm of tin (Sn). After validation, the model was used to investigate the effects of additional filtration of the incident beam with Pb and Sn. Supplementary incident x-ray spectra, representing heavier filtration (Pb: 2 and 3 mm; Sn: 1, 2, and 3 mm) were computationally generated and used with the model to obtain XRF/scatter spectra. Quasimonochromatic incident x-ray spectra (81, 85, 90, 95, and 100 keV with 10 keV full width at half maximum) were also investigated to determine the ideal energy for distinguishing gold XRF signal from the scatter background. Fluorescence signal-to-dose ratio (FSDR) and fluorescence-normalized scan time (FNST) were used as metrics to assess results. Calculated XRF/scatter spectra for 1-mm Pb and 0.9-mm Sn filters matched (r ≥ 0.996) experimental measurements. Calculated spectra representing additional filtration for both filter materials showed that the spectral hardening improved the FSDR at the expense of requiring a much longer FNST. In general, using Sn instead of Pb, at a given filter thickness, allowed an increase of up to 20% in FSDR, more prominent gold XRF peaks, and up to an order of magnitude decrease in FNST. Simulations using quasimonochromatic spectra suggested that increasing source x-ray energy, in the investigated range of 81-100 ke

  12. High resolution x-ray fluorescence spectroscopy - a new technique for site- and spin-selectivity

    International Nuclear Information System (INIS)

    Wang, Xin

    1996-12-01

    X-ray spectroscopy has long been used to elucidate electronic and structural information of molecules. One of the weaknesses of x-ray absorption is its sensitivity to all of the atoms of a particular element in a sample. Through out this thesis, a new technique for enhancing the site- and spin-selectivity of the x-ray absorption has been developed. By high resolution fluorescence detection, the chemical sensitivity of K emission spectra can be used to identify oxidation and spin states; it can also be used to facilitate site-selective X-ray Absorption Near Edge Structure (XANES) and site-selective Extended X-ray Absorption Fine Structure (EXAFS). The spin polarization in K fluorescence could be used to generate spin selective XANES or spin-polarized EXAFS, which provides a new measure of the spin density, or the nature of magnetic neighboring atoms. Finally, dramatic line-sharpening effects by the combination of absorption and emission processes allow observation of structure that is normally unobservable. All these unique characters can enormously simplify a complex x-ray spectrum. Applications of this novel technique have generated information from various transition-metal model compounds to metalloproteins. The absorption and emission spectra by high resolution fluorescence detection are interdependent. The ligand field multiplet model has been used for the analysis of Kα and Kβ emission spectra. First demonstration on different chemical states of Fe compounds has shown the applicability of site selectivity and spin polarization. Different interatomic distances of the same element in different chemical forms have been detected using site-selective EXAFS

  13. Full Field X-Ray Fluorescence Imaging Using Micro Pore Optics for Planetary Surface Exploration

    Science.gov (United States)

    Sarrazin, P.; Blake, D. F.; Gailhanou, M.; Walter, P.; Schyns, E.; Marchis, F.; Thompson, K.; Bristow, T.

    2016-01-01

    Many planetary surface processes leave evidence as small features in the sub-millimetre scale. Current planetary X-ray fluorescence spectrometers lack the spatial resolution to analyse such small features as they only provide global analyses of areas greater than 100 mm(exp 2). A micro-XRF spectrometer will be deployed on the NASA Mars 2020 rover to analyse spots as small as 120m. When using its line-scanning capacity combined to perpendicular scanning by the rover arm, elemental maps can be generated. We present a new instrument that provides full-field XRF imaging, alleviating the need for precise positioning and scanning mechanisms. The Mapping X-ray Fluorescence Spectrometer - "Map-X" - will allow elemental imaging with approximately 100µm spatial resolution and simultaneously provide elemental chemistry at the scale where many relict physical, chemical and biological features can be imaged in ancient rocks. The arm-mounted Map-X instrument is placed directly on the surface of an object and held in a fixed position during measurements. A 25x25 mm(exp 2) surface area is uniformly illuminated with X-rays or alpha-particles and gamma-rays. A novel Micro Pore Optic focusses a fraction of the emitted X-ray fluorescence onto a CCD operated at a few frames per second. On board processing allows measuring the energy and coordinates of each X-ray photon collected. Large sets of frames are reduced into 2d histograms used to compute higher level data products such as elemental maps and XRF spectra from selected regions of interest. XRF spectra are processed on the ground to further determine quantitative elemental compositions. The instrument development will be presented with an emphasis on the characterization and modelling of the X-ray focussing Micro Pore Optic. An outlook on possible alternative XRF imaging applications will be discussed.

  14. X-ray fluorescence analysis for trace element determination in foodstuff chemistry

    International Nuclear Information System (INIS)

    Wildanger, W.

    The physical fundamentals of X-ray fluorescence analysis are given and the routine spectrometers described. The basic principles are given of analytical methods used in qualitative and quantitative fluorescence analyses. Examples are given of the use of the method in a number of fields and the possibility and usefulness is discussed for the determination of trace elements in foodstuffs. The preparation of samples, preliminary concentration of components and calibration methods are discussed. (M.K.)

  15. Fluorescence imaging as a diagnostic of M-band x-ray drive condition in hohlraum with fluorescent Si targets

    International Nuclear Information System (INIS)

    Li, Qi; Hu, Zhimin; Yao, Li; Huang, Chengwu; Yuan, Zheng; Zhao, Yang; Xiong, Gang; Qing, Bo; Lv, Min; Zhu, Tuo; Deng, Bo; Li, Jin; Wei, Minxi; Zhan, Xiayu; Li, Jun; Yang, Yimeng; Su, Chunxiao; Yang, Guohong; Zhang, Jiyan; Li, Sanwei

    2017-01-01

    Fluorescence imaging of surrogate Si-doped CH targets has been used to provide a measurement for drive condition of high-energy x-ray (i.e. M-band x-ray) drive symmetry upon the capsule in hohlraum on Shenguang-II laser facility. A series of experiments dedicated to the study of photo-pumping and fluorescence effect in Si-plasma are presented. To investigate the feasibility of fluorescence imaging in Si-plasma, an silicon plasma in Si-foil target is pre-formed at ground state by the soft x-ray from a half-hohlraum, which is then photo-pumped by the K-shell lines from a spatially distinct laser-produced Si-plasma. The resonant Si photon pump is used to improve the fluorescence signal and cause visible image in the Si-foil. Preliminary fluorescence imaging of Si-ball target is performed in both Si-doped and pure Au hohlraum. The usual capsule at the center of the hohlraum is replaced with a solid Si-doped CH-ball (Si-ball). Since the fluorescence is proportional to the photon pump upon the Si-plasma, high-energy x-ray drive symmetry is equal to the fluorescence distribution of the Si-ball. (paper)

  16. X-ray fluorescence spectrometers: a comparison of wavelength and energy dispersive instruments

    International Nuclear Information System (INIS)

    Slates, R.V.

    1977-11-01

    Wavelength dispersive and energy dispersive x-ray fluorescence spectrometers are compared. Separate sections are devoted to principles of operation, sample excitation, spectral resolution, and x-ray detection. Tabulated data from the literature are cited in the comparison of accuracy, precision, and detection limits. Spectral interferences and distortions are discussed. Advantages and limitations are listed for simultaneous wavelength dispersive spectrometers, sequential wavelength dispersive spectrometers, and Si(Li) energy dispersive spectrometers. Accuracy, precision, and detection limits are generally superior for wavelength dispersive spectrometers

  17. SAVLOC, computer program for automatic control and analysis of X-ray fluorescence experiments

    Science.gov (United States)

    Leonard, R. F.

    1977-01-01

    A program for a PDP-15 computer is presented which provides for control and analysis of trace element determinations by using X-ray fluorescence. The program simultaneously handles data accumulation for one sample and analysis of data from previous samples. Data accumulation consists of sample changing, timing, and data storage. Analysis requires the locating of peaks in X-ray spectra, determination of intensities of peaks, identification of origins of peaks, and determination of a real density of the element responsible for each peak. The program may be run in either a manual (supervised) mode or an automatic (unsupervised) mode.

  18. Analysis of siliceous geologic materials by energy-dispersive X-ray fluorescence

    International Nuclear Information System (INIS)

    Roca, M.; Bayon, A.

    1987-01-01

    The determination of the elements Al, Si, K, Ca, Ti, Cr, Mn and Fe in siliceous geologic samples by energy-dispersive X-ray fluorescence is investigated using the most adequate excitation conditions: direct excitation mode (rhodium anode X-ray tube) for the former two elements, and the secondary targets titanium for K and Ca, and germanium for Ti, Cr, Mn and Fe. For the correction of matrix effects the use of ratio methods has been tested. Procedure files have been defined allowing the automatic simultaneous acquisition and processing of spectra. (author)

  19. Spatially resolved density and ionization measurements of shocked foams using x-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    MacDonald, M. J.; Keiter, P. A.; Montgomery, D. S.; Scott, H. A.; Biener, M. M.; Fein, J. R.; Fournier, K. B.; Gamboa, E. J.; Kemp, G. E.; Klein, S. R.; Kuranz, C. C.; LeFevre, H. J.; Manuel, M. J. -E.; Wan, W. C.; Drake, R. P.

    2016-09-28

    We present experiments at the Trident laser facility demonstrating the use of x-ray fluorescence (XRF) to simultaneously measure density, ionization state populations, and electron temperature in shocked foams. An imaging x-ray spectrometer obtained spatially resolved measurements of Ti K-α emission. Density profiles were measured from K-α intensity. Ti ionization state distributions and electron temperatures were inferred by fitting K-α spectra to spectra from CRETIN simulations. This work shows that XRF provides a powerful tool to complement other diagnostics to make equation of state measurements of shocked materials containing a suitable tracer element.

  20. Surface extended x-ray absorption fine structure of low-Z absorbates using fluorescence detection

    International Nuclear Information System (INIS)

    Stoehr, J.; Kollin, E.B.; Fischer, D.A.; Hastings, J.B.; Zaera, F.; Sette, F.

    1985-05-01

    Comparison of x-ray fluorescence yield (FY) and electron yield surface extended x-ray absorption fine structure spectra above the S K-edge for c(2 x 2) S on Ni(100) reveals an order of magnitude higher sensitivity of the FY technique. Using FY detection, thiophene (C 4 H 4 S) chemisorption on Ni(100) is studied with S coverages down to 0.08 monolayer. The molecule dissociates at temperatures as low as 100K by interaction with fourfold hollow Ni sites. Blocking of these sites by oxygen leaves the molecule intact

  1. X-Ray Fluorescence Spectrometry. II. Determination of Uranium in ores

    International Nuclear Information System (INIS)

    Bermudez Polonio, J.; Crus Castillo, F. de la; Fernandez Cellini, R.

    1961-01-01

    A method of analysis of uranium in ores by X-ray spectrometry was developed, using the internal standard technique. Strontium was found to be the most suitable internal standard for general use. A Norelco Philips X-ray fluorescent spectrometer was used in this work, equipped with a lithium fluoride crystal acting as a diffraction grating analyzer. The intensity of the uranium-L α 1 spectral line is calculated and related to corresponding strontium-K α spectral line, both detected with a Scintillation Counter. (Author) 31 refs

  2. Trends in hard X-ray fluorescence mapping: environmental applications in the age of fast detectors

    Energy Technology Data Exchange (ETDEWEB)

    Lombi, E.; Donner, E. [University of South Australia, Centre for Environmental Risk Assessment and Remediation, Mawson Lakes, South Australia (Australia); CRC CARE, PO Box 486, Salisbury, South Australia (Australia); Jonge, M.D. de; Paterson, D. [Australian Synchrotron, X-ray Fluorescence Microscopy, 800 Blackburn Road, Clayton, Victoria (Australia); Ryan, C.G. [CSIRO Earth Science and Resource Engineering, Normanby Road, Clayton, Victoria (Australia)

    2011-06-15

    Environmental samples are extremely diverse but share a tendency for heterogeneity and complexity. This heterogeneity poses methodological challenges when investigating biogeochemical processes. In recent years, the development of analytical tools capable of probing element distribution and speciation at the microscale have allowed this challenge to be addressed. Of these available tools, laterally resolved synchrotron techniques such as X-ray fluorescence mapping are key methods for the in situ investigation of micronutrients and inorganic contaminants in environmental samples. This article demonstrates how recent advances in X-ray fluorescence detector technology are bringing new possibilities to environmental research. Fast detectors are helping to circumvent major issues such as X-ray beam damage of hydrated samples, as dwell times during scanning are reduced. They are also helping to reduce temporal beamtime requirements, making particularly time-consuming techniques such as micro X-ray fluorescence ({mu}XRF) tomography increasingly feasible. This article focuses on {mu}XRF mapping of nutrients and metalloids in environmental samples, and suggests that the current divide between mapping and speciation techniques will be increasingly blurred by the development of combined approaches. (orig.)

  3. Determination of Fission Products in Irradiated Fuel by X-Ray Fluorescence

    DEFF Research Database (Denmark)

    Mogensen, Mogens Bjerg; Als-Nielsen, Jens Aage; Andersen, Niels Hessel

    X-ray fluorescence i s a well e s t a b l i s h e d analytical tool for measuring elemental composition of fairly large (~ 5 cm2) "cold" samples. A version of t h i s technique has been developed for a n a l y s i s of radial d i s t r i b u t i o n of f i s s i o n products Xe, Cs and Ba...... in the scanning d i rection and 2 mm long. The measured Xe Ko x-ray i n t e n s i t i e s are converted to absolute concentrations by comparing to the intens i t y from a Xe gas standard. In the case of Cs and Ba s o l id standards may be used. The X-ray fluorescence analysis i s compared to other techniques used...... to obtain radial f i s s i o n product profiles, i t i s shown how a combination of X-ray fluorescence and electron probe micro a n a l y s i s i s able to reveal the amount of Xe in the grain boundary porosities....

  4. Rapid limit tests for metal impurities in pharmaceutical materials by X-ray fluorescence spectroscopy using wavelet transform filtering.

    Science.gov (United States)

    Arzhantsev, Sergey; Li, Xiang; Kauffman, John F

    2011-02-01

    We introduce a new method for analysis of X-ray fluorescence (XRF) spectra based on continuous wavelet transform filters, and the method is applied to the determination of toxic metals in pharmaceutical materials using hand-held XRF spectrometers. The method uses the continuous wavelet transform to filter the signal and noise components of the spectrum. We present a limit test that compares the wavelet domain signal-to-noise ratios at the energies of the elements of interest to an empirically determined signal-to-noise decision threshold. The limit test is advantageous because it does not require the user to measure calibration samples prior to measurement, though system suitability tests are still recommended. The limit test was evaluated in a collaborative study that involved five different hand-held XRF spectrometers used by multiple analysts in six separate laboratories across the United States. In total, more than 1200 measurements were performed. The detection limits estimated for arsenic, lead, mercury, and chromium were 8, 14, 20, and 150 μg/g, respectively.

  5. X-ray fluorescence analysis of terbium oxide for rare earth impurities

    International Nuclear Information System (INIS)

    Chandola, L.C.; Machado, I.J.; Mohile, A.N.

    1975-01-01

    A method for the determination of Sm 2 O 3 , Eu 2 O 3 , Gd 2 O 3 , Dy 2 O 3 , Ho 2 O 3 and Y 2 O 3 in terbium oxide is described. The sample is converted to terbium oxalate, mixed with boric acid binder in the ratio 2:1, pelleted at a pressure of 20 tons over a boric acid backing pellet and irradiated with x-rays from a tungsten tube operated by Philips PW 1140 generator. The secondary x-rays thus generated are analysed by a LiF (200) crystal in Philips PW 1220 x-ray fluorescence spectrometer using suitable detectors. The minimum determination limit (MDL) is 0.01% for all rare earth oxides determined except for Y 2 O 3 for which it is 0.005%. (author)

  6. Optimizing Monocapillary Optics for Synchrotron X-ray Diffraction, Fluorescence Imaging, and Spectroscopy Applications

    International Nuclear Information System (INIS)

    Bilderback, Donald H.; Kazimirov, Alexander; Gillilan, Richard; Cornaby, Sterling; Woll, Arthur; Zha, Chang-Sheng; Huang Rong

    2007-01-01

    A number of synchrotron x-ray applications such as powder diffraction in diamond anvil cells, microbeam protein crystallography, x-ray fluorescence imaging, etc. can benefit from using hollow glass monocapillary optics to improve the flux per square micron on a sample. We currently draw glass tubing into the desired elliptical shape so that only one-bounce under total reflection conditions is needed to bring the x-ray beam to a focus at a 25 to 50 mm distance beyond the capillary tip. For modest focal spot sizes of 10 to 20 microns, we can increase the intensity per square micron by factors of 10 to 1000. We show some of the results obtained at CHESS and Hasylab with capillaries focusing 5 to 40 keV radiation, their properties, and how even better the experimental results could be if more ideal capillaries were fabricated in the future

  7. Determination of Ti, Cr, Cu and Ta in niobium oxide by X-ray fluorescence method

    International Nuclear Information System (INIS)

    Dixit, R.M.; Deshpande, S.S.

    1986-01-01

    An x-ray fluorescence method for the determination of Ti, Cr, Cu and Ta in niobium oxide has been developed. Samples/standards in powder form are mixed with boric acid in the proportion of 1:1 (400 mg. each). Double layer pellets are prepared by pressing this mixture over a primary boric acid pellet. Philips PW-1220, a semiautomatic x-ray spectrometer with tungsten target x-ray tube for excitation and LiF (200) crystal for dispersion have been used. The determination range is from 0.005 to 0.1per cent for Ti and Cr, 0.01 to 0.1per cent for Cu and 0.05 to 1per cent for Ta. (author)

  8. X-ray specular reflection and fluorescence study of nano-films

    International Nuclear Information System (INIS)

    Zheludeva, S.; Novikova, N.

    2001-01-01

    The techniques that combine the advantages of high-resolution structure sensitive x-ray methods with spectroscopic selectivity of data obtained are shown to be extremely promising for characterization of organic and inorganic nano films and nano structures. Fluorescence yield angular dependences exited by complicated evanescent wave / x-ray standing wave pattern at total reflection and glancing incidence can be used to detect structure position of different ions in organic systems and alien interfacial layers in inorganic multilayers;, to get information about interdiffusion at the interfaces of Langmuir- Blodgett (L-B) films and artificial inorganic - x-ray mirrors; to study ion permeation through L-B nano structures - models of biomembrans; to obtain nano - film thickness and density; to get precisely the parameters of small d-space multilayer mirrors, ets

  9. Dosage of silicon in a soluble silicate using an x-ray-fluorescence radioisotopic method

    International Nuclear Information System (INIS)

    Wasilewska, M.; Robert, A.

    1969-01-01

    A description is given of a spectrometer for X ray fluorescence analysis having a radio active excitation source. It has been applied to the analysis of the silicon contained in an industrial soluble silicate. A theoretical study has been made for this analysis of the operational conditions such as: the effect of the particle size, the dilution of the sample, the sensitivity as a function of the X ray excitation energy. It is possible to obtain a relative accuracy of 0,87 per cent for the silicon determination, for one standard deviation. A comparison is made of the sensitivity obtained using this apparatus for the Si determination with that which can be obtained using a conventional apparatus fitted with an X ray tube. (author) [fr

  10. Application of X-ray fluorescence analytical techniques in phytoremediation and plant biology studies

    International Nuclear Information System (INIS)

    Necemer, Marijan; Kump, Peter; Scancar, Janez; Jacimovic, Radojko; Simcic, Jurij; Pelicon, Primoz; Budnar, Milos; Jeran, Zvonka; Pongrac, Paula; Regvar, Marjana; Vogel-Mikus, Katarina

    2008-01-01

    Phytoremediation is an emerging technology that employs the use of higher plants for the clean-up of contaminated environments. Progress in the field is however handicapped by limited knowledge of the biological processes involved in plant metal uptake, translocation, tolerance and plant-microbe-soil interactions; therefore a better understanding of the basic biological mechanisms involved in plant/microbe/soil/contaminant interactions would allow further optimization of phytoremediation technologies. In view of the needs of global environmental protection, it is important that in phytoremediation and plant biology studies the analytical procedures for elemental determination in plant tissues and soil should be fast and cheap, with simple sample preparation, and of adequate accuracy and reproducibility. The aim of this study was therefore to present the main characteristics, sample preparation protocols and applications of X-ray fluorescence-based analytical techniques (energy dispersive X-ray fluorescence spectrometry-EDXRF, total reflection X-ray fluorescence spectrometry-TXRF and micro-proton induced X-ray emission-micro-PIXE). Element concentrations in plant leaves from metal polluted and non-polluted sites, as well as standard reference materials, were analyzed by the mentioned techniques, and additionally by instrumental neutron activation analysis (INAA) and atomic absorption spectrometry (AAS). The results were compared and critically evaluated in order to assess the performance and capability of X-ray fluorescence-based techniques in phytoremediation and plant biology studies. It is the EDXRF, which is recommended as suitable to be used in the analyses of a large number of samples, because it is multi-elemental, requires only simple preparation of sample material, and it is analytically comparable to the most frequently used instrumental chemical techniques. The TXRF is compatible to FAAS in sample preparation, but relative to AAS it is fast, sensitive and

  11. Fission times studies of the Z=124 superheavy nucleus by X-ray fluorescence

    International Nuclear Information System (INIS)

    Airiau, Maud

    2016-01-01

    Since the 1960's nuclear structure model have predicted the existence of an island of stability of superheavy elements. It should be located around the next magic numbers expected at N=172 or 184 and between Z=114 and 126 depending on the model. Very high fission barrier of a few MeV are predicted to be generated by microscopic effects for those nuclei for which large fission times distributions extended to very high fission times are induced. Fission time measurements of the superheavy element Z=124 have been made by us using the X-ray fluorescence technique, a method based on the filling of inner-shell electronic vacancies created during the collision leading to the formation of the compound nucleus. The aim of this experiment was to detect in coincidence both fission fragments and characteristic X-rays from the Z=124, created by the reaction 238 U+ 70,76 Ge. The main difficulty was to identify those X-rays due to the fact that gamma-rays from fission fragments were emitted in the same energy range, which affected our photon multiplicities for any fragment selection. This new difficulty brings an important limitation to the study of some particular superheavy elements by the X-ray fluorescence method. K X-rays spectra have been simulated using MCDF (Multi-Configuration-Dirac-Fock) and then compared to the experimental ones in order to get a maximal K X-ray multiplicity compatible with our data. The extracted results were about 6-7% for 76 Ge and from 12 to 14% for 70 Ge. Those values remain compatible with the experimental signature of long lifetime component observed for the same system but using a blocking technique in single crystals. (author) [fr

  12. Soft-x-ray fluorescence study of buried silicides in antiferromagnetically coupled Fe/Si multilayers

    Energy Technology Data Exchange (ETDEWEB)

    Carlisle, J.A.; Chaiken, A.; Michel, R.P. [Lawrence Berkeley National Lab., CA (United States)] [and others

    1997-04-01

    Multilayer films made by alternate deposition of two materials play an important role in electronic and optical devices such as quantum-well lasers and x-ray mirrors. In addition, novel phenomena like giant magnetoresistance and dimensional crossover in superconductors have emerged from studies of multilayers. While sophisticated x-ray techniques are widely used to study the morphology of multilayer films, progress in studying the electronic structure has been slower. The short mean-free path of low-energy electrons severely limits the usefulness of photoemission and related electron free path of low-energy electrons severely limit spectroscopies for multilayer studies. Soft x-ray fluorescence (SXF) is a bulk-sensitive photon-in, photon-out method to study valence band electronic states. Near-edge x-ray absorption fine-structure spectroscopy (NEXAFS) measured with partial photon yield can give complementary bulk-sensitive information about unoccupied states. Both these methods are element-specific since the incident x-ray photons excite electrons from core levels. By combining NEXAFS and SXF measurements on buried layers in multilayers and comparing these spectra to data on appropriate reference compounds, it is possible to obtain a detailed picture of the electronic structure. Results are presented for a study of a Fe/Si multilayer system.

  13. X-ray fluorescence from the element with atomic number Z=120

    International Nuclear Information System (INIS)

    Fregeau, M.O.; Morjean, M.; Bonnet, E.; Chbihi, A.; Frankland, J.D.; Jacquet, D.; Rivet, M.F.; Tassan-Got, L.; Dechery, F.; Drouart, A.; Nalpas, L.; Ledoux, X.; Parlog, M.; Parlog, M.; Ciortea, C.; Dumitriu, D.; Fluerasu, D.; Gugiu, M.; Gramegna, F.; Kravchuk, V.L.; Marchi, T.; Marchi, T.; Fabris, D.; Corsi, A.; Barlini, S.

    2012-01-01

    An atomic clock based on x-ray fluorescence yields has been used to estimate the mean characteristic time for fusion followed by fission in reactions 238 U + 64 Ni at 6.6 MeV/A. Inner shell vacancies are created during the collisions in the electronic structure of the possibly formed Z = 120 compound nuclei. The filling of these vacancies accompanied by a x-ray emission with energies characteristic of Z = 120 can take place only if the atomic transitions occur before nuclear fission. Therefore, the x-ray yield characteristic of the united atom with 120 protons is strongly related to the fission time and to the vacancy lifetimes. K x rays from the element with Z = 120 have been unambiguously identified from a coupled analysis of the involved nuclear reaction mechanisms and of the measured photon spectra. A minimum mean fission time τ f = 2.5 * 10 -18 s has been deduced for Z = 120 from the measured x-ray multiplicity. (authors)

  14. Sub-micron Hard X-ray Fluorescence Imaging of Synthetic Elements

    Science.gov (United States)

    Jensen, Mark P.; Aryal, Baikuntha P.; Gorman-Lewis, Drew; Paunesku, Tatjana; Lai, Barry; Vogt, Stefan; Woloschak, Gayle E.

    2013-01-01

    Synchrotron-based X-ray fluorescence microscopy (SXFM) using hard X-rays focused into sub-micron spots is a powerful technique for elemental quantification and mapping, as well as microspectroscopic measurement such as μ-XANES (X-ray absorption near edge structure). We have used SXFM to image and simultaneously quantify the transuranic element plutonium at the L3 or L2 edge as well as lighter biologically essential elements in individual rat pheochromocytoma (PC12) cells after exposure to the long-lived plutonium isotope 242Pu. Elemental maps reveal that plutonium localizes principally in the cytoplasm of the cells and avoids the cell nucleus, which is marked by the highest concentrations of phosphorus and zinc, under the conditions of our experiments. The minimum detection limit under typical acquisition conditions for an average 202 μm2 cell is 1.4 fg Pu/cell or 2.9 × 10−20 moles Pu/μm2, which is similar to the detection limit of K-edge SXFM of transition metals at 10 keV. Copper electron microscopy grids were used to avoid interference from gold X-ray emissions, but traces of strontium present in naturally occurring calcium can still interfere with plutonium detection using its Lα X-ray emission. PMID:22444530

  15. X-ray fluorescence analysis of low concentrations metals in geological samples and technological products

    Science.gov (United States)

    Lagoida, I. A.; Trushin, A. V.

    2016-02-01

    For the past several years many nuclear physics methods of quantitative elemental analysis have been designed. Many of these methods have applied in different devices which have become useful and effective instrument in many industrial laboratories. Methods of a matter structure analysis are based on the intensity detection of the X-ray radiation from the nuclei of elements which are excited by external X-ray source. The production of characteristic X-rays involves transitions of the orbital electrons of atoms in the target material between allowed orbits, or energy states, associated with ionization of the inner atomic shells. One of these methods is X-ray fluorescence analysis, which is widespread in metallurgical and processing industries and is used to identify and measure the concentration of the elements in ores and minerals on a conveyor belt. Samples of copper ore with known concentrations of elements, were taken from the Ural deposit. To excite the characteristic X-rays radionuclide sources 109Cd, with half-life 461.4 days were used. After finding the calibration coefficients, control measurements of samples and averaging of overall samples were made. The measurement error did not exceed 3%.

  16. Soft-x-ray fluorescence study of buried silicides in antiferromagnetically coupled Fe/Si multilayers

    International Nuclear Information System (INIS)

    Carlisle, J.A.; Chaiken, A.; Michel, R.P.

    1997-01-01

    Multilayer films made by alternate deposition of two materials play an important role in electronic and optical devices such as quantum-well lasers and x-ray mirrors. In addition, novel phenomena like giant magnetoresistance and dimensional crossover in superconductors have emerged from studies of multilayers. While sophisticated x-ray techniques are widely used to study the morphology of multilayer films, progress in studying the electronic structure has been slower. The short mean-free path of low-energy electrons severely limits the usefulness of photoemission and related electron free path of low-energy electrons severely limit spectroscopies for multilayer studies. Soft x-ray fluorescence (SXF) is a bulk-sensitive photon-in, photon-out method to study valence band electronic states. Near-edge x-ray absorption fine-structure spectroscopy (NEXAFS) measured with partial photon yield can give complementary bulk-sensitive information about unoccupied states. Both these methods are element-specific since the incident x-ray photons excite electrons from core levels. By combining NEXAFS and SXF measurements on buried layers in multilayers and comparing these spectra to data on appropriate reference compounds, it is possible to obtain a detailed picture of the electronic structure. Results are presented for a study of a Fe/Si multilayer system

  17. Development of confocal X-ray fluorescence (XRF) microscopy at the Cornell high energy synchrotron source

    International Nuclear Information System (INIS)

    Woll, A.R.; Huang, R.; Mass, J.; Bisulca, C.; Bilderback, D.H.; Gruner, S.; Gao, N.

    2006-01-01

    A confocal X-ray fluorescence microscope was built at the Cornell High Energy Synchrotron Source (CHESS) to obtain compositional depth profiles of historic paintings. The microscope consists of a single-bounce, borosilicate monocapillary optic to focus the incident beam onto the painting and a commercial borosilicate polycapillary lens to collect the fluorescent X-rays. The resolution of the microscope was measured by scanning a variety of thin metal films through this confocal volume while monitoring the fluorescence signal. The capabilities of the technique were then probed using test paint microstructures with up to four distinct layers, each having a thickness in the range of 10-80 microns. Results from confocal XRF were compared with those from stand-alone XRF and visible light microscopy of the paint cross-sections. A large area, high-resolution scanner is currently being built to perform 3D scans on moderately sized paintings. (orig.)

  18. Study of properties of chemically modified samples of halloysite mineral with X-ray fluorescence and X-ray powder diffraction methods

    International Nuclear Information System (INIS)

    Banaś, D.; Kubala-Kukuś, A.; Braziewicz, J.; Majewska, U.; Pajek, M.; Wudarczyk-Moćko, J.; Czech, K.; Garnuszek, M.; Słomkiewicz, P.; Szczepanik, B.

    2013-01-01

    Elemental and chemical composition of raw and activated samples of halloysite mineral using wavelength dispersive X-ray fluorescence (WDXRF), total reflection X-ray fluorescence (TXRF) and X-ray powder diffraction (XRPD) methods were determined. As the result, it has been shown that application of the complementary X-ray spectrometry techniques allows very precise observation of changes in composition of halloysite mineral samples caused by its chemical modifications. Sample preparation procedure and usability of the research methods applied are described in details. Procedure of activation of raw halloysite mineral samples by etching them in sulfuric acid of various concentrations has been described and discussed. The ability of the samples to adsorb lead from intentionally contaminated water was tested and confirmed. - Author-Highlights: • We measured elemental and chemical composition of raw and activated halloysite mineral samples. • We showed that X-ray techniques allow precise study of changes in the sample composition. • We describe procedure of activation of the samples by etching them in sulfuric acid. • We tested ability of halloysite mineral to absorb lead from contaminated water

  19. Characterization of energy response for photon-counting detectors using x-ray fluorescence

    International Nuclear Information System (INIS)

    Ding, Huanjun; Cho, Hyo-Min; Molloi, Sabee; Barber, William C.; Iwanczyk, Jan S.

    2014-01-01

    Purpose: To investigate the feasibility of characterizing a Si strip photon-counting detector using x-ray fluorescence. Methods: X-ray fluorescence was generated by using a pencil beam from a tungsten anode x-ray tube with 2 mm Al filtration. Spectra were acquired at 90° from the primary beam direction with an energy-resolved photon-counting detector based on an edge illuminated Si strip detector. The distances from the source to target and the target to detector were approximately 19 and 11 cm, respectively. Four different materials, containing silver (Ag), iodine (I), barium (Ba), and gadolinium (Gd), were placed in small plastic containers with a diameter of approximately 0.7 cm for x-ray fluorescence measurements. Linear regression analysis was performed to derive the gain and offset values for the correlation between the measured fluorescence peak center and the known fluorescence energies. The energy resolutions and charge-sharing fractions were also obtained from analytical fittings of the recorded fluorescence spectra. An analytical model, which employed four parameters that can be determined from the fluorescence calibration, was used to estimate the detector response function. Results: Strong fluorescence signals of all four target materials were recorded with the investigated geometry for the Si strip detector. The average gain and offset of all pixels for detector energy calibration were determined to be 6.95 mV/keV and −66.33 mV, respectively. The detector’s energy resolution remained at approximately 2.7 keV for low energies, and increased slightly at 45 keV. The average charge-sharing fraction was estimated to be 36% within the investigated energy range of 20–45 keV. The simulated detector output based on the proposed response function agreed well with the experimental measurement. Conclusions: The performance of a spectral imaging system using energy-resolved photon-counting detectors is very dependent on the energy calibration of the

  20. Study of uranium contamination of ground water in Punjab using X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Alrakabi, Muhanad; Singh, Gurjeet; Bhalla, Atul; Kumar, Sunil; Kumar, Sanjeev; Rai, Bimal; Singh, N.; Shahi, J.S.; Mehta, D.; Srivastava, Alok

    2010-01-01

    A number of reports have appeared in public media about uranium ingestion being a possible cause for cancer and increased birth rate abnormalities among children in the Malwa region of Punjab state in India. These reports link problems like cancer and Autism, with the presence of uranium in the ground waters of Malwa region. The concentration of uranium in drinking water from sources as varied as ground water, canal water supply and reverse osmosis system have been investigated using X-ray fluorescence technique. Samples from the thermal power plants in the regions and nearby ground waters were also analyzed to identify the source of contamination. The samples were collected with assistance of the officials from the Government of Punjab. More than half a litre of each of the water samples was dried at 60 deg-80 deg in an oven. Residue was collected using larger quantities of water samples in case of RO water samples. The elemental analysis of the residue was carried out using the Energy-Dispersive X-Ray Fluorescence (EDXRF) spectrometer consisting of an 42 Mo-anode X-ray tube (Panalytical, 2.5 kW) as an excitation source and a Si(Li) detector. A combination of selective absorbers of 30 Zn, 38 Sr, and 39 Y was used in the incident beam for improving the detection limit for Uranium by reducing the background and removing the 42 Mo K X-rays. The detection limit in ppb/litre depends upon the amount of residue

  1. Synchrotron radiation microbeam X-ray fluorescence analysis of zinc concentration in remineralized enamel in situ.

    Science.gov (United States)

    Matsunaga, Tsunenori; Ishizaki, Hidetaka; Tanabe, Shuji; Hayashi, Yoshihiko

    2009-05-01

    Remineralization is an indispensable phenomenon during the natural healing process of enamel decay. The incorporation of zinc (Zn) into enamel crystal could accelerate this remineralization. The present study was designed to investigate the concentration and distribution of Zn in remineralized enamel after gum chewing. The experiment was performed at the Photon Factory. Synchrotron radiation was monochromatized and X-rays were focused into a small beam spot. The X-ray fluorescence (XRF) from the sample was detected with a silicon (Si) (lithium (Li)) detector. X-ray beam energy was tuned to detect Zn. The examined samples were small enamel fragments remineralized after chewing calcium phosphate-containing gum in situ. The incorporation of Zn atom into hydroxyapatite (OHAP), the main component of enamel, was measured using Zn K-edge extended X-ray absorption fine structure (EXAFS) with fluorescence mode at the SPring-8. A high concentration of Zn was detected in a superficial area 10-microm deep of the sectioned enamel after gum chewing. This concentration increased over that in the intact enamel. The atomic distance between Zn and O in the enamel was calculated using the EXAFS data. The analyzed atomic distances between Zn and O in two sections were 0.237 and 0.240 nm. The present experiments suggest that Zn is effectively incorporated into remineralized enamel through the physiological processes of mineral deposition in the oral cavity through gum-chewing and that Zn substitution probably occurred at the calcium position in enamel hydroxyapatite.

  2. Multi-elemental analysis of marine sediments of Sorsogon Bay using x-ray fluorescence spectroscopy

    International Nuclear Information System (INIS)

    Gonzales, Ralph Roly A.; Quirit, Leni L.; Rosales, Colleen Marciel F.; Pabroa, Preciosa Corazon B.; Sta Maria, Efren J.

    2011-01-01

    Metal composition and nutrient loadings of our bodies of water, when uncontrolled, may cause harmful bacterial contamination and pose threats in aquatic and human life. Toxic and trace element inputs in Sorsogon Bay sediments were determined using nuclear analytical techniques, more specifically, x-ray fluorescence spectrometry, in this study. Pre-treated marine sediment samples from Sorsogon Bay were homogenized using SPEX # 8000 mixer/mill and agate mortar and pestle, pelletized into 31-mm flat discs using SPEX 3630 X-Press and analyzed using PAN Analytical Epsilon 5 EDX X-ray Fluorescence Spectrometer with the emission and transmission method using silver and germanium secondary targets. Spectrum fitting performed using AXIL (Analysis of X-ray Spectra by Iterative Least-Squares Fitting), a subprogram in Quantitative X-ray Analysis System developed by the International Atomic Energy Agency and Quantitative Analysis of Environmental Samples program, was used for quantification of results. Results indicate generally moderate to high metal enrichment, specifically manganese, lead, cadmium, zinc and copper. Mercury and iron level enrichment are found to be low, marking an improvement from previous studies indicating high enrichment of these metals. (author)

  3. Monitoring the mass of UF6 gas and uranium deposits in aluminium pipes using X-ray fluorescence and X-ray transmission gauges

    International Nuclear Information System (INIS)

    Packer, T.W.; Smith, S.M.

    1984-12-01

    In order to determine the enrichment of UF 6 gas in centrifuge plant pipework it is necessary to measure the mass of the gas (pressure) and the mass per unit area of any uranium deposited on the pipe. This paper shows that it is possible to determine the pressure of the UF 6 gas in pipes 120 mm in diameter using an energy-dispersive X-ray fluorescence spectrometer. Results are also given of transmission measurements made using a low power X-ray generator operated at two different applied voltages. A method of using the two measurements to determine the mass per unit area of deposited uranium is described. (author)

  4. A low cost multi-purpose experimental arrangement for variants in energy dispersive X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Nascimento Filho, V.F.; Silva, R.M.C.; Moraes, L.M.B.; Parreira, P.S.; Appoloni, R.C.; Silva, R.M.C.

    2005-01-01

    Based in an X-ray tower with four exits (two line and two point beams) experimental conditions were arranged to carry out variants in energy dispersive X-ray fluorescence analysis: (1) the conventional one (EDXRF), with excitation/detection of thin and thick samples, under vacuum and air atmosphere, (2) the X-ray energy dispersive micro- fluorescence analysis(μ-EDXRF), with 2D mapping, using a quartz capillar, (3) the total reflection X-ray fluorescence (TXRF), under He and air atmosphere, and (4) secondary target/polarized X-ray fluorescence (P-EDXRF). It was possible to use a Cu, Mo or W target on the X-ray tube, with or without filter (V, Fe, Ni and Zr), and Si(Li) or Si-PIN semicondutor detectors coupled to a multichannel analyzer. In addition, it was possible to use the point beam to carry out experiments on (5) X-ray radiography and (6) X-ray absorption, and the line beam on (7) X-ray diffractometry studies.

  5. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    Energy Technology Data Exchange (ETDEWEB)

    Wilson, Matthew D., E-mail: Matt.Wilson@stfc.ac.uk; Seller, Paul; Veale, Matthew C. [Science and Technology Facilities Council, Rutherford Appleton Laboratory, Harwell Campus,UK (United Kingdom); Connolley, Thomas [Diamond Light Source, I12 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal [Diamond Light Source, B16 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Grant, Patrick S.; Liotti, Enzo; Lui, Andrew [Department of Materials, University of Oxford Parks Road, Oxford (United Kingdom)

    2016-07-27

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm{sup 2} with one of the 80×80 pixels imaging an area equivalent to 13µm{sup 2}. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  6. Energetic electron processes fluorescence effects for structured nanoparticles X-ray analysis and nuclear medicine applications

    Energy Technology Data Exchange (ETDEWEB)

    Taborda, A.; Desbrée, A. [Institut de Radioprotection et de Sûreté Nucléaire (IRSN), PRP-HOM/SDI/LEDI, BP-17, 31, Avenue de la Division Leclerc, 92262 Fontenay-aux-Roses (France); Carvalho, A. [IEQUALTECS, Lda, Rua Dr. Francisco Sá Carneiro, 36, 2500-065 S. Gregório CLD (Portugal); Chaves, P.C. [C" 2TN, Campus Tecnológico e Nuclear, Instituto Superior Técnico, Universidade de Lisboa, EN10 km 139.7, 2685-066 Bobadela LRS (Portugal); Reis, M.A., E-mail: mareis@ctn.tecnico.ulisboa.pt [IEQUALTECS, Lda, Rua Dr. Francisco Sá Carneiro, 36, 2500-065 S. Gregório CLD (Portugal); C" 2TN, Campus Tecnológico e Nuclear, Instituto Superior Técnico, Universidade de Lisboa, EN10 km 139.7, 2685-066 Bobadela LRS (Portugal)

    2016-08-15

    Superparamagnetic iron oxide (SPIO) nanoparticles are widely used as contrast agents for nuclear magnetic resonance imaging (MRI), and can be modified for improved imaging or to become tissue-specific or even protein-specific. The knowledge of their detailed elemental composition characterisation and potential use in nuclear medicine applications, is, therefore, an important issue. X-ray fluorescence techniques such as particle induced X-ray emission (PIXE) or X-ray fluorescence spectrometry (XRF), can be used for elemental characterisation even in problematic situations where very little sample volume is available. Still, the fluorescence coefficient of Fe is such that, during the decay of the inner-shell ionised atomic structure, keV Auger electrons are produced in excess to X-rays. Since cross-sections for ionisation induced by keV electrons, for low atomic number atoms, are of the order of 10{sup 3} barn, care should be taken to account for possible fluorescence effects caused by Auger electrons, which may lead to the wrong quantification of elements having atomic number lower than the atomic number of Fe. Furthermore, the same electron processes will occur in iron oxide nanoparticles containing {sup 57}Co, which may be used for nuclear medicine therapy purposes. In the present work, simple approximation algorithms are proposed for the quantitative description of radiative and non-radiative processes associated with Auger electrons cascades. The effects on analytical processes and nuclear medicine applications are quantified for the case of iron oxide nanoparticles, by calculating both electron fluorescence emissions and energy deposition on cell tissues where the nanoparticles may be embedded.

  7. Combined X-ray fluorescence and absorption computed tomography using a synchrotron beam

    International Nuclear Information System (INIS)

    Hall, C

    2013-01-01

    X-ray computed tomography (CT) and fluorescence X-ray computed tomography (FXCT) using synchrotron sources are both useful tools in biomedical imaging research. Synchrotron CT (SRCT) in its various forms is considered an important technique for biomedical imaging since the phase coherence of SR beams can be exploited to obtain images with high contrast resolution. Using a synchrotron as the source for FXCT ensures a fluorescence signal that is optimally detectable by exploiting the beam monochromaticity and polarisation. The ability to combine these techniques so that SRCT and FXCT images are collected simultaneously, would bring distinct benefits to certain biomedical experiments. Simultaneous image acquisition would alleviate some of the registration difficulties which comes from collecting separate data, and it would provide increased information about the sample: functional X-ray images from the FXCT, with the morphological information from the SRCT. A method is presented for generating simultaneous SRCT and FXCT images. Proof of principle modelling has been used to show that it is possible to recover a fluorescence image of a point-like source from an SRCT apparatus by suitably modulating the illuminating planar X-ray beam. The projection image can be successfully used for reconstruction by removing the static modulation from the sinogram in the normal flat and dark field processing. Detection of the modulated fluorescence signal using an energy resolving detector allows the position of a fluorescent marker to be obtained using inverse reconstruction techniques. A discussion is made of particular reconstruction methods which might be applied by utilising both the CT and FXCT data.

  8. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    International Nuclear Information System (INIS)

    Wilson, Matthew D.; Seller, Paul; Veale, Matthew C.; Connolley, Thomas; Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal; Grant, Patrick S.; Liotti, Enzo; Lui, Andrew

    2016-01-01

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm"2 with one of the 80×80 pixels imaging an area equivalent to 13µm"2. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  9. Precision scan-imaging for paperboard quality inspection utilizing X-ray fluorescence

    Science.gov (United States)

    Norlin, B.; Reza, S.; Fröjdh, C.; Nordin, T.

    2018-01-01

    Paperboard is typically made up of a core of cellulose fibers [C6H10O5] and a coating layer of [CaCO3]. The uniformity of these layers is a critical parameter for the printing quality. Current quality control methods include chemistry based visual inspection methods as well as X-ray based methods to measure the coating thickness. In this work we combine the X-ray fluorescence signals from the Ca atoms (3.7 keV) in the coating and from a Cu target (8.0 keV) placed behind the paper to simultaneously measure both the coating and the fibers. Cu was selected as the target material since its fluorescence signal is well separated from the Ca signal while its fluorescence's still are absorbed sufficiently in the paper. A laboratory scale setup is built using stepper motors, a silicon drift detector based spectrometer and a collimated X-ray beam. The spectroscopic image is retrieved by scanning the paperboard surface and registering the fluorescence signals from Ca and Cu. The exposure time for this type of setups can be significantly improved by implementing spectroscopic imaging sensors. The material contents in the layers can then be retrieved from the absolute and relative intensities of these two signals.

  10. Determination of rare-earth elements in rocks by isotope-excited X-ray fluorescence spectrometry

    DEFF Research Database (Denmark)

    Kunzendorf, Helmar; Wollenberg, H.A.

    1970-01-01

    Isotope-excited X-ray fluorescence spectrometry furnishes a rapid determination of rare-earth elements in unprepared rock samples. The samples are excited by 241Am γ-rays, generating X-ray spectra on a multichannel pulse-height analyser. Gaussian peaks of the Kα and Kβ X-ray energies are treated ......-ray spectrometric scan of a longitudinally sliced drill core showed a close correlation between rare-earth abundances and appropriate minerals.......Isotope-excited X-ray fluorescence spectrometry furnishes a rapid determination of rare-earth elements in unprepared rock samples. The samples are excited by 241Am γ-rays, generating X-ray spectra on a multichannel pulse-height analyser. Gaussian peaks of the Kα and Kβ X-ray energies are treated...

  11. Determinations of elements in pepperbush standard reference material by neutron activation and X-ray fluorescence analyses

    International Nuclear Information System (INIS)

    Mizumoto, Yoshihiko; Okada, Takayuki; Tatsumi, Toshiya; Kusakabe, Toshio; Katsurayama, Kousuke; Iwata, Shiro.

    1988-01-01

    Elemental contents in Pepperbush standard reference material have been determined by neutron activation and X-ray fluorescence analyses. The standard samples of orchard leaves, tomato leaves, pine needles and Kale are used for the experiment. In the neutron activation analysis, gamma-ray spectra of nuclei produced by (n,γ) reaction on Pepperbush and standard samples are measured with Ge detectors. In the X-ray fluorescence analysis, the samples are excited with X-rays from X-ray tube with rhodium anode, and the characteristic X-rays from samples are measured with a proportional counter or NaI(Tl) detector. From the gamma- and X-ray intensities, the elemental contents in Pepperbush are determined. As a result, the contents of seventeen elements, such as sodium, calcium, iron, etc., in Pepperbush are determined. (author)

  12. Investigation of radiation absorption and X-ray fluorescence properties of medical imaging scintillators by Monte Carlo methods

    International Nuclear Information System (INIS)

    Nikolopoulos, D.; Kandarakis, I.; Cavouras, D.; Valais, I.; Linardatos, D.; Michail, C.; David, S.; Gaitanis, A.; Nomicos, C.; Louizi, A.

    2006-01-01

    X-ray absorption and X-ray fluorescence properties of medical imaging scintillating screens were studied by Monte Carlo methods as a function of the incident photon energy and screen-coating thickness. The scintillating materials examined were Gd 2 O 2 S (GOS) Gd 2 SiO 5 (GSO) YAlO 3 (YAP), Y 3 Al 5 O 12 (YAG), LuSiO 5 (LSO), LuAlO 3 (LuAP) and ZnS. Monoenergetic photon exposures were modeled in the range from 10 to 100 keV. The corresponding ranges of coating thicknesses of the investigated scintillating screens ranged up to 200 mg cm -2 . Results indicated that X-ray absorption and X-ray fluorescence are affected by the incident photon energy and the screen's coating thickness. Regarding incident photon energy, this X-ray absorption and fluorescence was found to exhibit very intense changes near the corresponding K edge of the heaviest element in the screen's scintillating material. Regarding coating thickness, thicker screens exhibited higher X-ray absorption and X-ray fluorescence. Results also indicated that a significant fraction of the generated X-ray fluorescent quanta escape from the scintillating screen. This fraction was found to increase with screen's coating thickness. At the energy range studied, most of the incident photons were found to be absorbed via one-hit photoelectric effect. As a result, the reabsorption of scattered radiation was found to be of rather minor importance; nevertheless this was found to increase with the screen's coating thickness. Differences in X-ray absorption and X-ray fluorescence were found among the various scintillators studied. LSO scintillator was found to be the most attractive material for use in many X-ray imaging applications, exhibiting the best absorption properties in the largest part of the energy range studied. Y-based scintillators were also found to be of significant absorption performance within the low energy ranges

  13. Probing symmetry and symmetry breaking in resonant soft-x-ray fluorescence spectra of molecules

    Energy Technology Data Exchange (ETDEWEB)

    Glans, P.; Gunnelin, K.; Guo, J. [Uppsala Univ. (Sweden)] [and others

    1997-04-01

    Conventional non-resonant soft X-ray emission brings about information about electronic structure through its symmetry and polarization selectivity, the character of which is governed by simple dipole rules. For centro-symmetric molecules with the emitting atom at the inversion center these rules lead to selective emission through the required parity change. For the more common classes of molecules which have lower symmetry or for systems with degenerate core orbitals (delocalized over identical sites), it is merely the local symmetry selectivity that provides a probe of the local atomic orbital contribution to the molecular orbital. For instance, in X-ray spectra of first row species the intensities essentially map the p-density at each particular atomic site, and, in a molecular orbital picture, the contribution of the local p-type atomic orbitals in the LCAO description of the molecular orbitals. The situation is different for resonant X-ray fluorescence spectra. Here strict parity and symmetry selectivity gives rise to a strong frequency dependence for all molecules with an element of symmetry. In addition to symmetry selectivity the strong frequency dependence of resonant X-ray emission is caused by the interplay between the shape of a narrow X-ray excitation energy function and the lifetime and vibrational broadenings of the resonantly excited core states. This interplay leads to various observable effects, such as linear dispersion, resonance narrowing and emission line (Stokes) doubling. Also from the point of view of polarization selectivity, the resonantly excited X-ray spectra are much more informative than the corresponding non-resonant spectra. Examples are presented for nitrogen, oxygen, and carbon dioxide molecules.

  14. Observation of X-ray shadings in synchrotron radiation-total reflection X-ray fluorescence using a color X-ray camera

    Energy Technology Data Exchange (ETDEWEB)

    Fittschen, Ursula Elisabeth Adriane, E-mail: ursula.fittschen@chemie.uni-hamburg.de [Institut für Anorganische und Angewandte Chemie, Universität Hamburg, Martin-Luther-King-Platz 6, 20146 Hamburg (Germany); Menzel, Magnus [Institut für Anorganische und Angewandte Chemie, Universität Hamburg, Martin-Luther-King-Platz 6, 20146 Hamburg (Germany); Scharf, Oliver [IfG Institute for Scientific Instruments GmbH, Berlin (Germany); Radtke, Martin; Reinholz, Uwe; Buzanich, Günther [BAM Federal Institute of Materials Research and Testing, Berlin (Germany); Lopez, Velma M.; McIntosh, Kathryn [Los Alamos National Laboratory, Los Alamos, NM (United States); Streli, Christina [Atominstitut, TU Wien, Vienna (Austria); Havrilla, George Joseph [Los Alamos National Laboratory, Los Alamos, NM (United States)

    2014-09-01

    Absorption effects and the impact of specimen shape on TXRF analysis has been discussed intensively. Model calculations indicated that ring shaped specimens should give better results in terms of higher counts per mass signals than filled rectangle or circle shaped specimens. One major reason for the difference in signal is shading effects. Full field micro-XRF with a color X-ray camera (CXC) was used to investigate shading, which occurs when working with small angles of excitation as in TXRF. The device allows monitoring the illuminated parts of the sample and the shaded parts at the same time. It is expected that sample material hit first by the primary beam shade material behind it. Using the CXC shading could be directly visualized for the high concentration specimens. In order to compare the experimental results with calculation of the shading effect the generation of controlled specimens is crucial. This was achieved by “drop on demand” technology. It allows generating uniform, microscopic deposits of elements. The experimentally measured shadings match well with those expected from calculation. - Highlights: • Use of a color X-ray camera and drop on demand printing to diagnose X-ray shading • Specimens were obtained uniform and well-defined in shape and concentration by printing. • Direct visualization and determination of shading in such specimens using the camera.

  15. Observation of X-ray shadings in synchrotron radiation-total reflection X-ray fluorescence using a color X-ray camera

    International Nuclear Information System (INIS)

    Fittschen, Ursula Elisabeth Adriane; Menzel, Magnus; Scharf, Oliver; Radtke, Martin; Reinholz, Uwe; Buzanich, Günther; Lopez, Velma M.; McIntosh, Kathryn; Streli, Christina; Havrilla, George Joseph

    2014-01-01

    Absorption effects and the impact of specimen shape on TXRF analysis has been discussed intensively. Model calculations indicated that ring shaped specimens should give better results in terms of higher counts per mass signals than filled rectangle or circle shaped specimens. One major reason for the difference in signal is shading effects. Full field micro-XRF with a color X-ray camera (CXC) was used to investigate shading, which occurs when working with small angles of excitation as in TXRF. The device allows monitoring the illuminated parts of the sample and the shaded parts at the same time. It is expected that sample material hit first by the primary beam shade material behind it. Using the CXC shading could be directly visualized for the high concentration specimens. In order to compare the experimental results with calculation of the shading effect the generation of controlled specimens is crucial. This was achieved by “drop on demand” technology. It allows generating uniform, microscopic deposits of elements. The experimentally measured shadings match well with those expected from calculation. - Highlights: • Use of a color X-ray camera and drop on demand printing to diagnose X-ray shading • Specimens were obtained uniform and well-defined in shape and concentration by printing. • Direct visualization and determination of shading in such specimens using the camera

  16. X-ray fluorescence/Auger-electron coincidence spectroscopy of vacancy cascades in atomic argon

    International Nuclear Information System (INIS)

    Arp, U.

    1996-01-01

    Argon L 2.3 -M 2.3 M 2.3 Auger-electron spectra were measured in coincidence with Kα fluorescent x-rays in studies of Ar K-shell vacancy decays at several photon energies above the K-threshold and on the 1s-4p resonance in atomic argon. The complex spectra recorded by conventional electron spectroscopy are greatly simplified when recorded in coincidence with fluorescent x-rays, allowing a more detailed analysis of the vacancy cascade process. The resulting coincidence spectra are compared with Hartree-Fock calculations which include shake-up transitions in the resonant case. Small energy shifts of the coincidence electron spectra are attributed to post-collision interaction with 1s photoelectrons

  17. Rapid analysis of molybdenum contents in molybdenum master alloys by X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Tongkong, P.

    1985-01-01

    Determination of molybdenum contents in molybdenum master alloy had been performed using energy dispersive x-ray fluorescence (EDX) technique where analysis were made via standard additions and calibration curves. Comparison of EDX technique with other analyzing techniques, i.e., wavelength dispersive x-ray fluorescence, neutron activation analysis and inductive coupled plasma spectrometry, showed consistency in the results. This technique was found to yield reliable results when molybdenum contents in master alloys were in the range of 13 to 50 percent using HPGe detector or proportional counter. When the required error was set at 1%, the minimum analyzing time was found to be 30 and 60 seconds for Fe-Mo master alloys with molybdenum content of 13.54 and 49.09 percent respectively. For Al-Mo master alloys, the minimum times required were 120 and 300 seconds with molybdenum content of 15.22 and 47.26 percent respectively

  18. Envelope method for background elimination from X-ray fluorescence spectra

    International Nuclear Information System (INIS)

    Monakhov, V.V.; Naumenko, P.A.; Chashinskaya, O.A.

    2006-01-01

    The influence of the background noise caused by Bremsstrahlung on the accuracy of the envelope method at x-ray fluorescence spectra processing is studied. This is carried out by the example of model spectra at different forms of Bremsstrahlung noise as well as at the presence of background noise in spectra. The interpolation by parabolic splines is used for the estimation of the error of the envelope method for the elimination of continuos background noise. It is found out that the error of the proposed method constitutes decimal parts of percent. It is shown that the envelope method is the effective technique for the elimination of the continuous Bremsstrahlung from x-ray fluorescence spectra of the first order [ru

  19. Nondestructive analysis of silver in gold foil using synchrotron radiation X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Kasamatsu, Masaaki; Suzuki, Yasuhiro; Suzuki, Shinichi; Nakanishi, Toshio; Shimoda, Osamu; Nishiwaki, Yoshinori; Miyamoto, Naoki

    2005-01-01

    Small particles of gold foil detached from an indoor decoration might be important evidence to associate a suspect with a crime scene. We have investigated the application of elemental analysis using synchrotron radiation X-ray fluorescence spectrometry to discriminate small particles of gold foil. Eight kinds of gold foil samples collected in Japan were used in the experiments. As a result of synchrotron radiation X-ray fluorescence spectrometry, only two elements, gold and silver, were detected from all gold foil samples. The intensity ratios of AgK α /AuL α showed good correlation with the content ratios of Ag/Au. The variation of intensity ratio within a same sample was sufficiently small compared with those of different samples. Therefore the comparison of this intensity ratio can be an effective method to discriminate small particles originating from different types of gold foil. (author)

  20. Trace elements determination in red and white wines using total-reflection X-ray fluorescence

    International Nuclear Information System (INIS)

    Anjos, M.J.; Lopes, R.T.; Jesus, E.F.O. de; Moreira, S.; Barroso, R.C.; Castro, C.R.F.

    2003-01-01

    Several wines produced in different regions from south of Brazil and available in markets in Rio de Janeiro were analyzed for their contents of elements such as: P, S, Cl, Ca, Ti, Cr, Mn, Fe, Ni, Cu, Zn, Rb and Sr. Multi-element analysis was possible with simple sample preparation and subsequent analysis by total-reflection X-ray fluorescence using synchrotron radiation. The measurement was carried at the X-ray fluorescence beamline in the Synchrotron Light Source Laboratory in Campinas, Brazil. The levels of the various elements obtained were lower in the Brazilian wines than the values generally found in the literature. The present study indicates the capability of multi-element analysis for determining the contents of various elements present in wines coming from Brazil vineyards by using a simple, sensitive and precise method

  1. Role of importance of X-ray fluorescence analysis of forensic samples

    International Nuclear Information System (INIS)

    Jha, Shailendra; Sharma, M.

    2009-01-01

    Full text: In the field of forensic science, it is very important to investigate the evidential samples obtained at various crime scenes. X-ray fluorescence (XRF) is used widely in forensic science [1]. Its main strength is its non-destructive nature, thus preserving evidence [2, 3]. In this paper, we report the application of XRF to examine the evidences like purity gold and silver jewelry (Indian Ornaments), remnants of glass pieces and paint chips recovered from crime scenes. The experimental measurements on these samples have been made using X-ray fluorescence spectrometer (LAB Center XRF-1800) procured from Shimazdu Scientific Inst., USA. The results are explained in terms of quantitative/ qualitative analysis of trace elements. (author)

  2. Single particle transfer for quantitative analysis with total-reflection X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Esaka, Fumitaka; Esaka, Konomi T.; Magara, Masaaki; Sakurai, Satoshi; Usuda, Shigekazu; Watanabe, Kazuo

    2006-01-01

    The technique of single particle transfer was applied to quantitative analysis with total-reflection X-ray fluorescence (TXRF) spectrometry. The technique was evaluated by performing quantitative analysis of individual Cu particles with diameters between 3.9 and 13.2 μm. The direct quantitative analysis of the Cu particle transferred onto a Si carrier gave a discrepancy between measured and calculated Cu amounts due to the absorption effects of incident and fluorescent X-rays within the particle. By the correction for the absorption effects, the Cu amounts in individual particles could be determined with the deviation within 10.5%. When the Cu particles were dissolved with HNO 3 solution prior to the TXRF analysis, the deviation was improved to be within 3.8%. In this case, no correction for the absorption effects was needed for quantification

  3. Beer analysis by synchrotron radiation total reflection X-ray fluorescence (SR-TXRF)

    International Nuclear Information System (INIS)

    Moreira, Silvana; Vives, Ana Elisa S. de; Nascimento Filho, Virgilio F.; Zucchi, Orgheda L.D.A.

    2005-01-01

    In this work the concentrations of P, S, Cl, K, Ca, Mn, Fe, Zn and Br in twenty-nine brands of national and international beers were determined by Synchrotron Radiation Total Reflection X-Ray Fluorescence analysis (SR-TXRF). The results were compared with the limits established by the Brazilian Legislation and the nutritive values established by National Agricultural Library (NAL). The measurements were performed at the X-ray Fluorescence Beamline at Synchrotron Light Source Laboratory, in Campinas, Sao Paulo, Brazil, using a polychromatic beam for excitation. A small volume of 5 μL of sample beers containing just an internal standard, used to correct geometry effects, were analyzed without any pre-treatment. The measuring time was 100 s and the detection limits obtained varied from 1μg.L -1 for Mn and Fe to 15μg.L -1 for P. (author)

  4. Determination of copper in geological materials by X-ray fluorescence

    International Nuclear Information System (INIS)

    Roca, M.; Bayon, A.

    1981-01-01

    X-ray fluorescence has been applied to the determination of copper content of geological materials in the concentration range of 0.01 to % CuO. A molybdenum target tube Is used, samples being presented in finely-ground powder form. Various methods for the correction for background and Instrumental copper interferences have been considered. To correct for matrix effects different tube scattered primary radiations have been tested as references or internal standards. MoK(41 - (C) provides the most suitable results. The use of influence empirical coefficients for the effect of iron on copper and of mass absorption coefficients has also been considered. For samples with a high content of lead, several procedures to correct for I t s influence have been investigated. Comparison between data obtained by X-ray fluorescence and wet-chemical techniques indicated good agreement. (Author) 6 refs

  5. Multielement characterization of atmospheric pollutants by x-ray fluorescence analysis and instrumental neutron activation analysis

    International Nuclear Information System (INIS)

    Rancitelli, L.A.; Tanner, T.M.

    1976-01-01

    The simultaneous measurement of a wide spectrum of elements in aerosols collected on air filters and in rainwater can yield information on the origin, transport, and removal of atmospheric pollutants. In order to determine the elemental content of these aerosols, a pair of highly sensitive, precise and complementing instrumental techniques, x-ray fluorescence and neutron activation analysis, have been developed and employed. Data are presented on the results of combined x-ray fluorescence and activation analysis of aerosols collected in a number of urban areas of the USA and from the 80th median sampling network in March 1972. From a comparison of these ratios in granite and diabase with those of filters placed in urban areas, it is evident that Zn, Se, Sb, Hg, and Pb levels have been increased by as much as several orders of magnitude. Al, Co, La, Fe, Eu, Sm, Tb, Ta, Hf, and Th appear to exist at levels compatible with an earth's crust origin

  6. Determination of small amounts of aluminum in iron and steel by x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Ito, Minao; Sato, Shoki; Narita, Masanao

    1981-01-01

    The accuracy of X-ray fluorescence analysis for determining aluminum in steel is generally poor when the aluminum content is below 0.1%. In this study, it is confirmed that the existence of inhomogeneous acid-insoluble aluminum causes the poor accuracy of X-ray fluorescence analysis. The accuracy can be improved by adding granular ferro-zirconium to molten steel when sampling. By using this method, the accuracy of determining aluminum was 0.0024% for low alloy steels and 0.0022% for stainless steels as the standard deviation obtained for a series of differences between an analytical result and a standard value. Corrections for overlapping of the chromium spectrum, and for absorption and enhancement effects of co-existent elements are not necessary for the analysis of low alloy steels, whereas these corrections are necessary for the analysis of stainless steels. (author)

  7. Application of the Total Reflection X-ray Fluorescence technique to trace elements determination in tobacco

    International Nuclear Information System (INIS)

    Martinez, T.; Lartigue, J.; Zarazua, G.; Avila-Perez, P.; Navarrete, M.; Tejeda, S.

    2008-01-01

    Many studies have identified an important number of toxic elements along with organic carcinogen molecules and radioactive isotopes in tobacco. In this work we have analyzed by Total Reflection X-Ray Fluorescence 9 brands of cigarettes being manufactured and distributed in the Mexican market. Two National Institute of Standards and Technology standards and a blank were equally treated at the same time. Results show the presence of some toxic elements such as Pb and Ni. These results are compared with available data for some foreign brands, while their implications for health are discussed. It can be confirmed that the Total Reflection X-Ray Fluorescence method provides precise (reproducible) and accuracy (trueness) data for 15 elements concentration in tobacco samples

  8. Sample analysis using gamma ray induced fluorescent X-ray emission

    Energy Technology Data Exchange (ETDEWEB)

    Sood, B S; Allawadhi, K L; Gandhi, R; Batra, O P; Singh, N [Punjabi Univ., Patiala (India). Nuclear Science Labs.

    1983-01-01

    A non-destructive method for the analysis of materials using gamma ray-induced fluorescent x-ray emission has been developed. In this method, special preparation of very thin samples in which the absorption of the incident gamma rays and the emitted fluorescent x-rays is negligible, is not needed, and the absorption correction is determined experimentally. A suitable choice of the incident gamma ray energies is made to minimise enhancement effects through selective photoionization of the elements in the sample. The method is applied to the analysis of a typical sample of the soldering material using 279 keV and 59.5 keV gamma rays from /sup 203/Hg and /sup 241/Am radioactive sources respectively. The results of the analysis are found to agree well with those obtained from the chemical analysis.

  9. Application of the Total Reflection X-ray Fluorescence technique to trace elements determination in tobacco

    Science.gov (United States)

    Martinez, T.; Lartigue, J.; Zarazua, G.; Avila-Perez, P.; Navarrete, M.; Tejeda, S.

    2008-12-01

    Many studies have identified an important number of toxic elements along with organic carcinogen molecules and radioactive isotopes in tobacco. In this work we have analyzed by Total Reflection X-Ray Fluorescence 9 brands of cigarettes being manufactured and distributed in the Mexican market. Two National Institute of Standards and Technology standards and a blank were equally treated at the same time. Results show the presence of some toxic elements such as Pb and Ni. These results are compared with available data for some foreign brands, while their implications for health are discussed. It can be confirmed that the Total Reflection X-Ray Fluorescence method provides precise (reproducible) and accuracy (trueness) data for 15 elements concentration in tobacco samples.

  10. Beer analysis by synchrotron radiation total reflection X-ray fluorescence (SR-TXRF)

    Energy Technology Data Exchange (ETDEWEB)

    Moreira, Silvana [Universidade Estadual de Campinas, SP (Brazil). Faculdade de Engenharia Civil, Arquitetura e Urbanismo. Dept. de Recursos Hidricos]. E-mail: silvana@fec.unicamp.br; Vives, Ana Elisa S. de [Universidade Metodista de Piracicaba (UNIMEP), Santa Barbara D' Oeste, SP (Brazil). Faculdade de Engenharia, Arquitetura e Urbanismo]. E-mail: aesvives@unimep.br; Nascimento Filho, Virgilio F. [Centro de Energia Nuclear na Agricultura (CENA), Piracicaba, SP (Brazil). Lab. de Instrumentacao Nuclear]. E-mail: virgilio@cena.usp.br; Zucchi, Orgheda L.D.A. [Sao Paulo Univ., Ribeirao Preto, SP (Brazil). Faculdade de Ciencias Farmaceuticas de Ribeirao Preto]. E-mail: olzucchi@fcfrp.usp.br

    2005-07-01

    In this work the concentrations of P, S, Cl, K, Ca, Mn, Fe, Zn and Br in twenty-nine brands of national and international beers were determined by Synchrotron Radiation Total Reflection X-Ray Fluorescence analysis (SR-TXRF). The results were compared with the limits established by the Brazilian Legislation and the nutritive values established by National Agricultural Library (NAL). The measurements were performed at the X-ray Fluorescence Beamline at Synchrotron Light Source Laboratory, in Campinas, Sao Paulo, Brazil, using a polychromatic beam for excitation. A small volume of 5 {mu}L of sample beers containing just an internal standard, used to correct geometry effects, were analyzed without any pre-treatment. The measuring time was 100 s and the detection limits obtained varied from 1{mu}g.L{sup -1} for Mn and Fe to 15{mu}g.L{sup -1} for P. (author)

  11. Quantitative analysis by X-ray fluorescence using first principles for matrix correction

    International Nuclear Information System (INIS)

    Hulett, L.D.; Dunn, H.W.; Tarter, J.G.

    1978-01-01

    The quantitative interpretation of X-ray fluorescence (XRF) data is often difficult because of matrix effects. The intensity of fluorescence measured for a given element is not only dependent on the element's concentration, but also on the mass absorption coefficients of the sample for the excitation and fluorescence radiation. Also, there are interelement effects in which high-energy fluorescence from heavier elements is absorbed by lighter elements with a resulting enhancement of their fluorescence. Recent theoretical treatments of this problem have shown that X-ray fluorescence data can be corrected for these matrix effects by calculations based on first principles. Fundamental constants, available in atomic physics data tables, are the only parameters needed. It is not necessary to make empirical calibrations. The application of this correctional procedure to alloys and alumina-supported catalysts is described. A description is given of a low-background spectrometer which uses monochromatic Ag Ksub(α) radiation for excitation. Matrix corrections by first principles can be easily applied to data from instruments of this type because fluorescence excitation cross-sections and mass absorption coefficients can be accurately defined for monochromatic radiation. (author)

  12. Determination of rare-earth elements in rocks by isotope-excited X-ray fluorescence spectrometry

    DEFF Research Database (Denmark)

    Kunzendorf, Helmar; Wollenberg, H.A.

    1970-01-01

    Isotope-excited X-ray fluorescence spectrometry furnishes a rapid determination of rare-earth elements in unprepared rock samples. The samples are excited by 241Am γ-rays, generating X-ray spectra on a multichannel pulse-height analyser. Gaussian peaks of the Kα and Kβ X-ray energies are treated......-ray spectrometric scan of a longitudinally sliced drill core showed a close correlation between rare-earth abundances and appropriate minerals....

  13. A set-up of micro-X-ray fluorescence system based on polycapillary X-ray optics and applications for archaeology

    International Nuclear Information System (INIS)

    Cheng Lin; Pan Qiuli; Ding Xunliang; Liu Zhiguo

    2008-01-01

    The paper concerns in the structures, performances and characteristics and applications for archaeology of a new micro-X-ray fluorescence system based on rotating anode X-ray generator and polycapillary X-ray optics. The polycapillary X-ray optics used here can focus the primary X-ray beam down to some tens of micrometers in diameters that allows for non-destructive and local analysis of sub-mm samples with minor/ trace level sensitivity. In order to prove the potentials of this instrument used in archaeology, a piece of Chinese ancient blue and white porcelain produced in Ming Dynasty was analyzed. The results show that intensities of Mn-Kα, Co-Kα are variable in agree with the thick of blue glaze. The correlation analysis indicates the Mn and Co have the best correlations. So, the concentrations or ratios of Mn and Co are crucial to determine the provenance and identify from a fake one of Chinese ancient blue and white porcelain. (authors)

  14. Non-scanning x-ray fluorescence microscope: application to real time micro-imaging

    International Nuclear Information System (INIS)

    Sakurai, K.; Eba, H.

    2000-01-01

    So far, x-ray fluorescence (XRF) micro-imaging has been performed by a 2D positional scan of a sample against a collimated beam. Obtaining information on specific elements in a nondestructive manner is an attractive prospect for many scientific applications. Furthermore, a synchrotron micro-beam can enhance the spatial resolution down to 0.1 μm. However, the total measuring time becomes quite long (a few hours to a half day), since one needs a number of scanning points in order to obtain a high-quality image. It is possible to obtain an x-ray image with 1 M pixels and with 20 μm resolution in a very short time of 20 sec - 3 min using a non-scanning XRF microscope, which is based on completely different concept. In the present report, we discuss the application of this technique to real time micro-imaging. The experiments were carried out at BL-4A, Photon Factory, Tsukuba, Japan. We employed a grazing-incidence arrangement to make primary x-rays illuminate the whole sample surface. We adopted parallel-beam optics and extremely-close-geometry in order to detect x-ray fluorescence with a CCD camera. The selective-excitation capability of tunable monochromatic synchrotron radiation is a feasible method for distinguishing the elements of interest. One can obtain an image of each element by differentiating the images obtained above and below the absorption edges of interest. The growth of metallic dendrites from a solution dropped on a substrate was studied successfully. Several different growth patterns, corresponding to concentration and other conditions for diffusion, were observed as x-ray images. Since the present technique requires only 40 sec for each shot, it is possible to record a growing process through repeated exposures like a movie. The authors would like to thank Prof. A. Iida (Photon Factory) for his valuable comments. (author)

  15. A gas microstrip X-ray detector for soft energy fluorescence EXAFS

    CERN Document Server

    Smith, A D; Derbyshire, G E; Duxbury, D M; Lipp, J; Spill, E J; Stephenson, R

    2001-01-01

    Gas microstrip detectors have been previously developed by the particle physics community, where their robustness, compactness and high counting speed have been recognised. These features are particularly attractive to synchrotron radiation use. In this paper, we describe a gas microstrip detector employing multi-element readout and specifically developed for high count rate fluorescence EXAFS at soft X-ray energies below 4 keV.

  16. Determination of cesium and selenium in cultivated mushrooms using radionuclide X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Racz, L.; Bumbalova, A.; Harangozo, M.; Toelgyessy, J.; Tomecek, O.

    2000-01-01

    Cesium and selenium intake of cultivated mushrooms (Agaricus bisporus), with these elements previously added to culture medium, has been examined from the viewpoint of health- and environmental protection. The process of measuring has been carried out by the radionuclide X-ray fluorescence technique. Treatments of the elementary substance with Se salt appears to influence the Se content of the mushrooms to a significant extent. Cs intake is of considerable importance, as this element is accumulated by mushrooms. (author)

  17. Determination of low concentrations of thorium in geological materials by X-ray fluorescence

    International Nuclear Information System (INIS)

    Roca, M.; Bayon, A.

    1981-01-01

    An X-ray fluorescence method for the determination of thorium in geological samples down to 2 ppm ThO 2 has been developed. To achieve this determination limit an exposed area of the sample 42.5 mm in diameter is used, working with a molybdenum target tube operated at 90 kV and 30 m A. Corrections for background and line interference of the Rb Kα radiation have been carefully considered and empirical correction coefficients calculated. (Author) 3 refs

  18. Applications of total reflection X-ray fluorescence in multi-element analysis

    International Nuclear Information System (INIS)

    Michaelis, W.; Prange, A.; Knoth, J.

    1985-01-01

    Although Total Reflection X-Ray Fluorescence Analysis (TXRF) became available for practical applications and routine measurements only few years ago, the number of programmes that make use of this method is increasing rapidly. The scope of work is widespread over environmental research and monitoring, mineralogy, mineral exploration, oceanography, biology, medicine and biochemistry. The present paper gives a brief survey of these applications and summarizes some of them which are typical for quite different matrices. (orig.)

  19. Synchrotron radiation total reflection X-ray fluorescence (SR-TXRF) for evaluation of food contamination

    International Nuclear Information System (INIS)

    Vives, A.E.S.; Brienza, S.M.B.; Moreira, S.; Zucchi, O.L.A.; Nascimento Filho, V.F.

    2006-01-01

    The objective of this study was to use synchrotron radiation total reflection X-ray fluorescence analysis (SR-TXRF) for the determination of metals and other elements in food available to the population in commercial establishments, in order to evaluate the risks of contamination by these products. The analyzed species were vegetables, leafy vegetables, fruits, cereal and grain. The results indicated that some species were contaminated by Cr, Cu, Zn and Pb with concentrations much higher than the reference values. (author)

  20. Zinc content determination in rice and other agriproducts by X-ray fluorescence

    International Nuclear Information System (INIS)

    Denis Alpizar, Otoniel; Diaz Rizo, Oscar

    2009-01-01

    The Zn content in regular consumption foodstuff (rice, some vegetables and roots) in Cuba is reported. Concentrations are determined by X-Ray Fluorescence analysis using a set of organic standards doped with Zn. The accuracy of the analytical procedure was validated using the Certified Reference Materials IAEA 393 and y MA-B-3/TM. The obtained results show rice as the major Zn bioaccumulator of the studied agriproducts and the main Zn source in Cuban human diet. (author)

  1. Zinc content determination in rice and other agricultural products by X-ray fluorescence

    International Nuclear Information System (INIS)

    Denis Alpizar, Otoniel; Diaz Rizo, Oscar

    2010-01-01

    The Zn content in regular consumption foodstuff (rice, some vegetables and roots) in Cuba is reported. Concentrations are determined by X-Ray Fluorescence analysis using a set of organic standards doped with Zn. The accuracy of the analytical procedure was validated using the Certified Reference Materials IAEA 393 and y MA-B-3/TM. The obtained results show rice as the major Zn bioaccumulator of the studied agricultural products and the main Zn source in Cuban human diet. (author)

  2. Radioisotope induced X-ray fluorescence analysis of cereal grains and flour

    International Nuclear Information System (INIS)

    Toeroek, Sz.; Szoekefalvi-Nagy, Z.

    1982-06-01

    Radioisotope-induced X-ray fluorescence analysis is a rather simple and easy method for investigating ashed plant material. In order to reduce matrix effects thin samples of 2 mg/cm 2 are analysed to obtain a reasonable compromise between maximum sensitivity and the lowest possible absorption effects. Concentrations are determined by standard addition method. An accuracy of 6-8% can be achieved. As an application analytical results are given for whole grains of several sorts of wheat. (author)

  3. Analysis of trace elements in medicinal plants with energy dispersive X-ray fluorescence

    International Nuclear Information System (INIS)

    Ekinci, N.; Polat, R.; Budak, G.; Ekinci, R.

    2004-01-01

    Mankind still depend on traditional herbal medicine for the treatment of various diseases and ailments. Elemental composition and concentration of medicinal plants have been investigated by energy dispersive X-ray fluorescence. The elements present in medicinal plants are P, Cl, K, Ca, S, Al, Ti, V, Rb, Sr, Zr, Nb, Mo, In, Sn, I and Ce. The physical basis of the used analytical method, the experimental set up and the procedure of sample preparation are presented. (author)

  4. Criteria to stablish an absolute semiquantitative analytic method by x-ray fluorescence

    International Nuclear Information System (INIS)

    Echeverria, Fernando.

    1990-01-01

    This work had been developed by fluorescence x-ray laboratory of Nuclear Studies Centre form Ecuador, it pretent to stablish an appropiate discernement of a FRX spectrum to know the concentration range form sample elements, minerals principaly, submit to a FRX. The study tries to stablish the best way to calibration the instruments like the adecuate samples preparation to do the analysis by FRX

  5. Quantitative determination of uranium in organic solution by X-ray fluorescence

    International Nuclear Information System (INIS)

    Leyt, D.V. de; Colangelo, C.H.

    1987-01-01

    An X-ray fluorescent method for the determination of uranium in tributilphosphate-kerosene-nitriacid solution has been developed. Chemical properties of the matrix elements were studied in order to select a convenient procedure to determine samples and standards on the same way. The method avoids the destruction of the organic material and has proved to be very useful for the fast control of uranium concentration. (Author) [es

  6. Development of unified X-ray fluorescent analysis to determine rhenium content in multicomponent oxide compositions

    International Nuclear Information System (INIS)

    Drobot, D.V.; Belyaev, A.V.; Kutvitskij, V.A.; Rysev, A.P.

    1999-01-01

    A procedure to prepare rhenium-containing glass-like specimens on the basis of bismuth and boron oxides is proposed. The glasses produced are studied by X-ray fluorescent analysis and routine spectrometric thiocyanate analysis. The results make it possible to determine rhenium in oxide mixtures in the range of its content 0.01 - 10% with S r = 0.03 [ru

  7. Ferro-metry and X-ray fluorescence to measure wear

    International Nuclear Information System (INIS)

    Catherin, J.Y.

    1997-01-01

    The determination of wear of metallic structures is carried out by measuring metallic particles in the lubricating oil. This method is routinely used by SNCF, the French railroad company to control the motors of the high speed trains (TGV). Four methods, plasma emission spectrometry, X-ray fluorescence analysis, ferro-metry and magnetometry are used. The field of application and advantage of each method is described. (C.B.)

  8. X-ray fluorescence microscopy reveals the role of selenium in spermatogenesis

    OpenAIRE

    Kehr, Sebastian; Malinouski, Mikalai; Finney, Lydia; Vogt, Stefan; Labunskyy, Vyacheslav M.; Kasaikina, Marina V.; Carlson, Bradley A.; Zhou, You; Hatfield, Dolph L.; Gladyshev, Vadim N.

    2009-01-01

    Selenium (Se) is a trace element with important roles in human health. Several selenoproteins have essential functions in development. However, the cellular and tissue distribution of Se remains largely unknown because of the lack of analytical techniques that image this element with sufficient sensitivity and resolution. Herein, we report that X-ray fluorescence microscopy (XFM) can be used to visualize and quantify the tissue, cellular and subcellular topography of Se. We applied this techn...

  9. X-ray fluorescence analysis of metal concentration in an alloy electroplating bath

    International Nuclear Information System (INIS)

    Hines, R.A.

    1980-06-01

    An energy-dispersive x-ray fluorescence analysis system has been developed for rapid, simultaneous analysis of gold and copper concentrations in an aqueous electroplating bath. The speed and repeatability of the system make it well suited for in-process control. Data collection and reduction are automatic. The analysis requires less than 10 minutes from taking the sample to printing the gold and copper concentrations

  10. 3D Synchrotron μ-x-ray fluorescence analysis on human bones

    International Nuclear Information System (INIS)

    Zoeger, N.; Wobrauschek, P.; Streli, C.; Chinea-Cano, E.; Wegrzynek, D.; Roschger, P.; Simon, R.; Staub, S.; Falkenberg, G.

    2004-01-01

    A comparison between μ-x-ray fluorescence tomography and confocal μ-x-ray fluorescence analysis (μ-XRF) will be presented. These techniques were used to study the three dimensional (3D) elemental distribution in human bone. Since bone shows very strong inhomogeneities in structure as well as in distribution of the chemical elements, two dimensional (2D) analysis (element mapping) of the samples always led to difficulties in interpreting the results and assigning elemental distributions to microscopic structures. Tomography scans in fluorescence and absorption mode have been carried out simultaneously at the fluo-topo beamline at ANKA, Karlsruhe, to determine the distribution of the elements over the depth of the previously prepared sample from human patella. A monochromatized x-ray beam (17 keV) from a bending magnet station focused by a compound refractive lens to a beamsize of 10 x 5 μm was used to perform the measurements. The transmitted beam signal measured with the SD detector was utilized to apply a simplified absorption correction to XRF tomographic images. Based on the XRF sinograms the elemental distribution within the object cross-section was reconstructed by means of filtered backprojection. The same section of human bone has been analyzed by confocal μ-XRF at HASYLAB, Hamburg, Germany beamline L. With this experiment two polycapillary half lenses were used; one for focusing the previously monochromatized primary x-ray beam onto the sample and the second half lens in front of a Si(Li) detector to get a small inspected area. By overlapping the two foci of the lenses a very well defined volume of investigation could be defined. Scanning the sample up- and downstream it was possible to determine the elemental distribution in depth of the sample. An absorption correction has been applied to get a corrected fluorescence image of the sample. Both methods showed consistent results and allowed a precise localization of the elements of interest. (author)

  11. Energy-dispersive X-ray fluorescence analysis of cerium in ferrosilicon

    International Nuclear Information System (INIS)

    Marbec, E.R.

    1987-01-01

    The cerium was determined in ferrosilicon samples by energy-dispersive X-ray fluorescence techniques (XRF) techniques, with a secondary target of gadolinium. The methods employed were: comparison and linear regression with reference materials with cerium concentration between 0.4 and 1.0%. The samples were prepared in the form of pellets and the analytical results are reported as an average of five determinations with a confidence limits at 95% probability. (Author) [es

  12. Determination of calcium and iron in limestone by X-ray fluorescence method

    International Nuclear Information System (INIS)

    Sovtsova, M.K.

    1977-01-01

    The results of determining calcium and iron content in limestone by X-ray fluorescence method are described. The 109 Cd isotape was chosen as a source for excitation, as it permited to reduce the concentration degeneration in the range of large Ca contents due to the larger energy of the primary radiation. The root-mean-square deviation from the data of chemical analysis was +-0.02%FeO and +-0.22%CaO

  13. Multielemental analysis of surface sediments in Havana bay (Cuba) using X-ray fluorescence

    International Nuclear Information System (INIS)

    Gelen, A.; Corrales, Y.; Lopez, N.; Manso Guevara, M. V.; Casanova, A. O.; Alessandro, K. D'; Diaz, O.; Espen, P. Van; Beltran, J.; Soto, J.

    2006-01-01

    Multielemental Analysis was performed in Superficial Sediments in Havana Bay. Twenty one samples were analysed by Dispersive Energy X- Ray Fluorescence using an spectrometer based on Si (Li) semiconductor detector an a 109 Cd source. The results showed a similar behaviour in the levels of contamination related with neutron activation analysis. The data suggest that an anthropogenic input into the bay from domestic sewage and industries occurred. (Full text)

  14. Analysis of noble metal on automotive exhaust catalysts by radioisotope-induce x-ray fluorescence

    International Nuclear Information System (INIS)

    Elgart, M.F.

    1976-01-01

    A technique was developed for the in-situ analysis of noble metals deposited on monolithic automotive exhaust catalysts. This technique is based on radioisotope-induced x-ray fluorescence, and provides a detailed picture of the distribution of palladium and platinum on catalyst samples. The experimental results for the cross section of a monolithic exhaust catalyst, analyzed in increments of 0.2 cm 3 , are compared with analyses for palladium and platinum obtained by instrumental neutron activation analysis

  15. Determination of potassium concentration in organic samples by means of x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Soto Moran, R.L.; Szgedi, S.

    1993-01-01

    By means of x-ray fluorescence analysis and the inner standard method using KH 2 PO 4 as the added chemical compound, potassium concentration of roots, stems , leaf, flowers and grains from Quinua (Chenopodium Quinoa Willd). which was previously treated with a nitrogen ed fertilizers has been determined taking into account the increasing effect the average atomic number due to used standard. Experimental errors are lower than 10 %

  16. Total reflection X-ray fluorescence analysis with synchrotron radiation monochromatized by multilayer structures

    International Nuclear Information System (INIS)

    Rieder, R.; Wobrauschek, P.; Ladisich, W.; Streli, C.; Aiginger, H.; Garbe, S.; Gaul, G.; Knoechel, A.; Lechtenberg, F.

    1995-01-01

    To achieve lowest detection limits in total reflection X-ray fluorescence analysis (TXRF) synchrotron radiation has been monochromatized by a multilayer structure to obtain a relative broad energy band compared to Bragg single crystals for an efficient excitation. The energy has been set to 14 keV, 17.5 keV, 31 keV and about 55 keV. Detection limits of 20 fg and 150 fg have been achieved for Sr and Cd, respectively. ((orig.))

  17. Application of instrumental neutron activation analysis and X-ray fluorescence analysis in art pieces investigation

    International Nuclear Information System (INIS)

    Panczyk, E.; Kierzek, J.; Walis, L.; Ligeza, M.

    1996-01-01

    The application of instrumental neutron activation analysis have been shown for the trace element identification in dyes of old painting and other art objects. The recognition of their composition is a important measure for attribution. Also the X-ray fluorescence analysis has been frequently used for examination of art objects. The age determination of the old chinese porcelain is a good example described in the paper. 20 refs, 4 figs

  18. OI Fluorescent Line Contamination in Soft X-Ray Diffuse Background Obtained with Suzaku/XIS

    OpenAIRE

    Sekiya, Norio; Yamasaki, Noriko Y.; Mitsuda, Kazuhisa; Takei, Yoh

    2014-01-01

    The quantitative measurement of OVII line intensity is a powerful method for understanding the soft X-ray diffuse background. By systematically analyzing the OVII line intensity in 145 high-latitude Suzaku/XIS observations, the flux of OI fluorescent line in the XIS spectrum, contaminating the OVII line, is found to have an increasing trend with time especially after 2011. For these observations, the OVII line intensity would be overestimated unless taking into consideration the OI fluorescen...

  19. New x-ray optical system for fluorescence beamline at Hasylab

    International Nuclear Information System (INIS)

    Falkenberg, G.; Tschentscher, T.

    2000-01-01

    Beamline L at HASYLAB/DESY is actually dedicated to micro x-ray fluorescence (μ-XRF) experiments using the white beam from a bending magnet of the storage ring DORIS III. To extend the applicability of beamline L to other x-ray fluorescence techniques, such as synchrotron radiation total reflection x-ray fluorescence (SR-TXRF) and micro x-ray absorption near edge structures in fluorescence mode (μ-XANES), new x-ray optics have been designed and are under installation at the moment. The suitability of beamline L for SR-TXRF experiments has been shown previously in a number of studies using temporary setups for beam monochromatization and collimation. The new optical system comprises a slit system, a pair of x-ray mirrors for focussing, collimation and high energy cut-off (12 keV and 30 keV), a double multilayer monochromator for broad bandpass applications (TXRF) and a double perfect-crystal monochromator for spectroscopy (XANES, speciation). The multilayer monochromator will utilize a pair of NiC with a spacing of 4.0 nm for the energy range 2-10 keV and a second pair of WB 4 C with a spacing of 3.0 nm for the range 4-30 keV. To extend the energy range for broad bandpass applications to higher photon energies SiGe gradient crystals are foreseen (ΔE/E ∼ 10 -3 ). For the perfect-crystal monochromator we have chosen a pair of Ge 111 crystals for the energy range 2-10 keV and Si 111 crystals for 7-90 keV. To enable the use of low photon energies down to 2 keV the monochromator vessel is sealed to the ring vacuum by a 25 μm thick carbon window. The mirrors and monochromators deflect the beam vertically and can be moved out of the beam independently. Fixed exit geometry permits the illumination of the same sample spot with different wavelength and energy bands. All optical elements accept the full vertical beam opening in order to enable both vertical and horizontal geometries for sample and detector. (author)

  20. X-ray fluorescence analysis of ancient and medieval brass artifacts from south Moravia

    International Nuclear Information System (INIS)

    Hložek, M.; Komoróczy, B.; Trojek, T.

    2012-01-01

    This paper deals with an investigation of archeological finds using X-ray fluorescence analysis and microanalysis. The main aim of the investigation was to prove the production of brass in the South Moravian Region (part of the Czech Republic) in former times. The probable brass production technology is described. Various objects dating back to Antiquity and to the Middle Ages were investigated using two X-ray fluorescence systems, and the results of the analyses are discussed. The measurements showed, e.g., that fragments of Roman scale armor and a belt fitting dating back to Antiquity were made of brass. Brass was also identified on the surfaces of various ancient and medieval molds and melting pots. - Highlights: ► Semiquantitative X-ray fluorescence analysis of archeological finds. ► Two different gilding techniques of a brass belt terminal found in Brno. ► Use of brass before the Great Moravian period. ► Evidence of brass casting in the 12th century in Brno.

  1. Kα resonance fluorescence in Al, Ti, Cu and potential applications for X-ray sources

    Science.gov (United States)

    Nahar, Sultana N.; Pradhan, Anil K.

    2015-04-01

    The Kα resonance fluorescence (RFL) effect via photoabsorptions of inner shell electrons as the element goes through multiple ionization states is studied. We demonstrate that the resonances observed recently in Kα (1s-2p) fluorescence in aluminum plasmas by using a high-intensity X-ray free-electron laser [1] are basically K-shell resonances in hollow atoms going through multiple ionization states at resonant energies as predicted earlier for gold and iron ions [2]. These resonances are formed below the K-shell ionization edge and shift toward higher energies with ionization states, as observed. Fluorescence emission intensities depend on transition probabilities for each ionization stage of the given element for all possible Kα (1 s → 2 p) transition arrays. The present calculations for resonant photoabsorptions of Kα photons in Al have reproduced experimentally observed features. Resonant cross sections and absorption coefficients are presented for possible observation of Kα RFL in the resonant energy ranges of 4.5-5.0 keV for Ti ions and 8.0-8.7 keV for Cu ions respectively. We suggest that theoretically the Kα RFL process may be driven to enhance the Auger cycle by a twin-beam monochromatic X-ray source, tuned to the K-edge and Kα energies, with potential applications such as the development of narrow-band biomedical X-ray devices.

  2. X-ray fluorescence analysis of ancient and medieval brass artifacts from south Moravia

    Energy Technology Data Exchange (ETDEWEB)

    Hlozek, M. [Methodical Centre of Conservation-Technical Museum in Brno, Purkynova 105, 612 00 Brno (Czech Republic); Komoroczy, B. [Institute of Archeology of the Academy of Science of the Czech Republic, Kralovopolska 147, 612 00 Brno (Czech Republic); Trojek, T., E-mail: tomas.trojek@fjfi.cvut.cz [Department of Dosimetry and Application of Ionizing Radiation, Czech Technical University in Prague, Brehova 7, 115 19 Praha 1 (Czech Republic)

    2012-07-15

    This paper deals with an investigation of archeological finds using X-ray fluorescence analysis and microanalysis. The main aim of the investigation was to prove the production of brass in the South Moravian Region (part of the Czech Republic) in former times. The probable brass production technology is described. Various objects dating back to Antiquity and to the Middle Ages were investigated using two X-ray fluorescence systems, and the results of the analyses are discussed. The measurements showed, e.g., that fragments of Roman scale armor and a belt fitting dating back to Antiquity were made of brass. Brass was also identified on the surfaces of various ancient and medieval molds and melting pots. - Highlights: Black-Right-Pointing-Pointer Semiquantitative X-ray fluorescence analysis of archeological finds. Black-Right-Pointing-Pointer Two different gilding techniques of a brass belt terminal found in Brno. Black-Right-Pointing-Pointer Use of brass before the Great Moravian period. Black-Right-Pointing-Pointer Evidence of brass casting in the 12th century in Brno.

  3. A method of measuring gold nanoparticle concentrations by x-ray fluorescence for biomedical applications

    Energy Technology Data Exchange (ETDEWEB)

    Wu Di; Li Yuhua; Wong, Molly D.; Liu Hong [Center for Bioengineering and School of Electrical and Computer Engineering, University of Oklahoma, Norman, Oklahoma 73019 (United States)

    2013-05-15

    Purpose: This paper reports a technique that enables the quantitative determination of the concentration of gold nanoparticles (GNPs) through the accurate detection of their fluorescence radiation in the diagnostic x-ray spectrum. Methods: Experimentally, x-ray fluorescence spectra of 1.9 and 15 nm GNP solutions are measured using an x-ray spectrometer, individually and within chicken breast tissue samples. An optimal combination of excitation and emission filters is determined to segregate the fluorescence spectra at 66.99 and 68.80 keV from the background scattering. A roadmap method is developed that subtracts the scattered radiation (acquired before the insertion of GNP solutions) from the signal radiation acquired after the GNP solutions are inserted. Results: The methods effectively minimize the background scattering in the spectrum measurements, showing linear relationships between GNP solutions from 0.1% to 10% weight concentration and from 0.1% to 1.0% weight concentration inside a chicken breast tissue sample. Conclusions: The investigation demonstrated the potential of imaging gold nanoparticles quantitatively in vivo for in-tissue studies, but future studies will be needed to investigate the ability to apply this method to clinical applications.

  4. Development of suitable plastic standards for X-ray fluorescence analysis

    Energy Technology Data Exchange (ETDEWEB)

    Mans, Christian [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: c.mans@fh-muenster.de; Hanning, Stephanie [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: hanning@fh-muenster.de; Simons, Christoph [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: simons@fh-muenster.de; Wegner, Anne [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: awegner@fh-muenster.de; Janssen, Anton [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: janssena@fh-muenster.de; Kreyenschmidt, Martin [University of Applied Sciences Muenster, Department of Chemical Engineering, Advanced Analytical Chemistry, Stegerwaldstr. 39, 48565 Steinfurt (Germany)], E-mail: martin.kreyenschmidt@fh-muenster.de

    2007-02-15

    For the adoption of the EU directive 'Restriction on use of certain Hazardous Substances' and 'Waste Electrical and Electronic Equipment' using X-ray fluorescence analysis suitable standard materials are required. Plastic standards based on acrylonitrile-butadiene-styrene terpolymer, containing the regulated elements Br, Cd, Cr, Hg and Pb were developed and produced as granulates and solid bodies. The calibration materials were not generated as a dilution from one master batch but rather the element concentrations were distributed over nine independent calibration samples. This was necessary to enable inter-elemental corrections and empirical constant mass absorption coefficients. The produced standard materials are characterized by a homogenous element distribution, which is more than sufficient for X-ray fluorescence analysis. Concentrations for all elements except for Br could be determined by Inductively Coupled Plasma Atomic Emission Spectroscopy after microwave assisted digestion. The concentration of Br was determined by use of Neutron Activation Analysis at Hahn-Meitner-Institute in Berlin, Germany. The correlation of the X-ray fluorescence analysis measurements with the values determined using Inductively Coupled Plasma Atomic Emission Spectroscopy and Neutron Activation Analysis showed a very good linearity.

  5. Chromosome translocations in chinese medical X-ray workers analyzed by fluorescence in situ hybridization

    International Nuclear Information System (INIS)

    Sun Yuanming; Li Jin; Wang Qin; Tang Weisheng; Wang Zhiquan

    2002-01-01

    Objective: To study long-term radiation effect in occupational workers exposed to low dose X-rays using the method of fluorescence in situ hybridization (FISH). Method: Chromosome translocations of 25 medical X-ray workers were analyzed by FISH with chromosome No. 4 and No. 7 probes according to PAINT (The Protocol for Aberration Identification and Nomenclature Terminology) system. Results: The frequency of genome translocation in X-ray workers was (13.14 ± 1.23)/1000 cells. The rate of complete and incomplete translocation was 1:1.7. According to the calendar year of entry before/after the year of 1965 as the border, the data showed that the incomplete translocation of the after 1965 group was obviously higher than those of the controls (P < 0.01 and P < 0.05, respectively). Conclusion: The chromosome translocation in early Chinese medical X-ray workers is mainly the incomplete one, the frequency of translocation does not dependent on chromosomal DNA content, and incomplete and complete ones increase along with prolongation of working years in their position

  6. Analysis of metals in organic compounds by energy dispersive X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Anjos, Marcelino J.; Lopes, Ricardo T.; Jesus, Edgar F.O. de

    2000-01-01

    Using energy dispersive X-ray fluorescence analysis with an X-ray tube filtered with Ti. It was possible to determine the concentration of the elements at ppm level of several elements: K, Ca, Ti, Cr, Mn, Fe, Ni, Cu, Zn As, Rb, Sr, Y, Zr, and Pb in two types of organic compound enough used in the agriculture: organic compound of urban garbage (Fertilurb) and aviary bed (birds manure). The experimental setup is composed of: x-ray tube (Oxford, 30 kV, 50 μA and W anode), an ORTEC Si-Li detector, with an energy resolution of about 180 eV at 5.9 keV and an ORTEC multichannel-analyser. The X-ray beam is quasi- monochromatic by using Ti filter. The samples were prepared in pellet form with superficial density in the range of 100 mg/cm 2 . The fundamental parameter method was used in order to verify the elemental concentration. The radiation transmission method was going used to the radiation absorption effects correction in the samples. (author)

  7. LabVIEW control software for scanning micro-beam X-ray fluorescence spectrometer.

    Science.gov (United States)

    Wrobel, Pawel; Czyzycki, Mateusz; Furman, Leszek; Kolasinski, Krzysztof; Lankosz, Marek; Mrenca, Alina; Samek, Lucyna; Wegrzynek, Dariusz

    2012-05-15

    Confocal micro-beam X-ray fluorescence microscope was constructed. The system was assembled from commercially available components - a low power X-ray tube source, polycapillary X-ray optics and silicon drift detector - controlled by an in-house developed LabVIEW software. A video camera coupled to optical microscope was utilized to display the area excited by X-ray beam. The camera image calibration and scan area definition software were also based entirely on LabVIEW code. Presently, the main area of application of the newly constructed spectrometer is 2-dimensional mapping of element distribution in environmental, biological and geological samples with micrometer spatial resolution. The hardware and the developed software can already handle volumetric 3-D confocal scans. In this work, a front panel graphical user interface as well as communication protocols between hardware components were described. Two applications of the spectrometer, to homogeneity testing of titanium layers and to imaging of various types of grains in air particulate matter collected on membrane filters, were presented. Copyright © 2012 Elsevier B.V. All rights reserved.

  8. X-ray optics for scanning fluorescence microscopy and other applications

    International Nuclear Information System (INIS)

    Ryon, R.W.; Warburton, W.K.

    1992-05-01

    Scanning x-ray fluorescence microscopy is analogous to scanning electron microscopy. Maps of chemical element distribution are produced by scanning with a very small x-ray beam. Goal is to perform such scanning microscopy with resolution in the range of <1 to 10 μm, using standard laboratory x-ray tubes. We are investigating mirror optics in the Kirkpatrick-Baez (K-B) configuration. K-B optics uses two curved mirrors mounted orthogonally along the optical axis. The first mirror provides vertical focus, the second mirror provides horizontal focus. We have used two types of mirrors: synthetic multilayers and crystals. Multilayer mirrors are used with lower energy radiation such as Cu Kα. At higher energies such as Ag Kα, silicon wafers are used in order to increase the incidence angles and thereby the photon collection efficiency. In order to increase the surface area of multilayers which reflects x-rays at the Bragg angle, we have designed mirrors with the spacing between layers graded along the optic axis in order to compensate for the changing angle of incidence. Likewise, to achieve a large reflecting surface with silicon, the wafers are placed on a specially designed lever arm which is bent into a log spiral by applying force at one end. In this way, the same diffracting angle is maintained over the entire surface of the wafer, providing a large solid angle for photon collection

  9. Determinations of silicon and phosphorus in Pepperbush standard reference material by neutron activation and x-ray fluorescence methods

    International Nuclear Information System (INIS)

    Mizumoto, Yoshihiko; Nishio, Hirofumi; Hayashi, Takeshi; Kusakabe, Toshio; Iwata, Shiro.

    1987-01-01

    Silicon and phosphorus contents in Pepperbush standard reference material were determined by neutron activation and X-ray fluorescence methods. In neutron activation analysis, β-ray spectra of 32 P produced by 31 P(n,γ) 32 P reaction on Pepperbush and standard samples were measured by a low background β-ray spectrometer. In X-ray fluorescence analysis, the standard samples were prepared by mixing the Pepperbush powder with silicon dioxide and diammonium hydrogenphosphate. Characteristic X-rays from the samples were analyzed by a wavelength dispersive X-ray fluorescence spectrometer. From the β and X-ray intensities, silicon and phosphorus contents in Pepperbush were determined to be 1840 ± 80 and 1200 ± 50 μg g -1 , respectively. (author)

  10. Application of the X-ray fluorescence analysis and X-ray diffraction in geochemical studies of the Pleistocene tills from Holy Cross Mountains

    International Nuclear Information System (INIS)

    Kubala-Kukuś, A.; Ludwikowska-Kedzia, M.; Banaś, D.; Braziewicz, J.; Majewska, U.; Pajek, M.; Wudarczyk-Moćko, J.

    2013-01-01

    X-ray fluorescence analysis methods (wavelength dispersive X-ray fluorescence analysis (WDXRF) and total reflection X-ray fluorescence (TXRF)) and X-ray powder diffraction (XRPD) have been applied in complementary geochemical studies of the Pleistocene till samples. The XRPD technique gave information about the mineral composition of the analyzed samples while the WDXRF and TXRF studies allowed the fast elemental analysis. The till samples were collected from different regions of Holy Cross Mountains (located in central Poland) which are still not unambiguously described in the context of the geochemical studies of the Quaternary sediments. The analysis was concentrated on the geochemical composition of the till samples both for materials occurring on the surface (characterized by continuous weathering processes) and for samples taken from core borehole. The overriding purpose of these studies is determination of the local lithotype of the tills and its lithologic and petrographic diagnostic properties, including the chemical composition of clay and minerals found in the clay. In the presented work the experimental sets up, sample preparation procedure and measurements programme will be discussed in details. Finally, the elemental and mineral compositions will be presented for studied different groups of the samples. - Highlights: • XRF analysis and X-ray diffraction used in studies of the till samples. • The till samples were collected from different regions of Holy Cross Mountains. • The analysis concentrates both on the samples from surface and from core borehole. • The purpose is determination of the local lithotype of the tills. • The experimental setup, sample preparation, measurements and results are discussed

  11. Use of planar HPGe detector as a part of X-ray fluorescent spectrometer for educational purposes

    International Nuclear Information System (INIS)

    Verenchikova, M.S.; Kalinin, V.N.; Mikhajlov, V.A.

    2011-01-01

    This work shows the possibility of use of the nondedicated gamma and X-ray detection head on the basis of planar HPGe detector with a big sensitive area equal to 2000 mm''2 as a part of X-ray fluorescent spectrometer during students' practicum.

  12. Simulation study of two-energy X-ray fluorescence holograms reconstruction algorithm to remove twin images

    International Nuclear Information System (INIS)

    Xie Honglan; Hu Wen; Luo Hongxin; Deng Biao; Du Guohao; Xue Yanling; Chen Rongchang; Shi Shaomeng; Xiao Tiqiao

    2008-01-01

    Unlike traditional outside-source holography, X-ray fluorescence holography is carded out with fluorescent atoms in a sample as source light for holographic imaging. With the method, three-dimensional arrangement of atoms into crystals can be observed obviously. However, just like traditional outside-source holography, X-ray fluorescence holography suffers from the inherent twin-image problem, too. With a 27-Fe-atoms cubic lattice as model, we discuss in this paper influence of the photon energy of incident source in removing twin images in reconstructed atomic images by numerical simulation and reconstruction with two-energy X-ray fluorescence holography. The results indicate that incident X-rays of nearer energies have better effect of removing twin images. In the detector of X-ray holography, minimum difference of the two incident energies depends on energy resolution of the monochromator and detector, and for inside source X-ray holography, minimum difference of the two incident energies depends on difference of two neighboring fluorescent energies emitting from the element and energy resolution of detector. The spatial resolution of atomic images increases with the incident energies. This is important for experiments of X-ray fluorescence holography, which is being developed on Shanghai Synchrotron Radiation Facility. (authors)

  13. Characterization of uranium in bituminized radioactive waste drums by self-induced X-ray fluorescence

    International Nuclear Information System (INIS)

    Pin, Patrick; Perot, Bertrand

    2013-06-01

    This paper reports the experimental qualification of an original uranium characterization method based on fluorescence X rays induced by the spontaneous gamma emission of bituminized radioactive waste drums. The main 661.7 keV gamma ray following the 137 Cs decay produces by Compton scattering in the bituminized matrix an intense photon continuum around 100 keV, i.e. in the uranium X-ray fluorescence region. 'Self-induced' X-rays produced without using an external source allow a quantitative assessment of uranium as 137 Cs and uranium are homogeneously mixed and distributed in the bituminized matrix. The paper presents the experimental qualification of the method with real waste drums, showing a detection limit well below 1 kg of uranium in 20 min acquisitions while the usual gamma rays of 235 U (185 keV) or 238 U (1001 keV of 234m Pa in the radioactive decay chain) are not detected. The relative uncertainty on the uranium mass assessed by self-induced X-ray fluorescence (SXRF) is about 50%, with a 95% confidence level, taking into account the correction of photon attenuation in the waste matrix. This last indeed contains high atomic numbers elements like uranium, but also barium, in quantities which are not known for each drum. Attenuation is estimated thanks to the peak-to-Compton ratio to limit the corresponding uncertainty. The SXRF uranium masses measured in the real drums are in good agreement with long gamma-ray spectroscopy measurements (1001 keV peak) or with radiochemical analyses. (authors)

  14. Feasibility study for the in vivo measurement of lead in bone using L-x-ray fluorescence

    International Nuclear Information System (INIS)

    Wielopolski, L.; Slatkin, D.N.; Vartsky, D.; Ellis, K.J.; Cohn, S.H.

    1980-01-01

    Lead deposits in bone were detected by x-ray fluorescence using x-rays from either a 125 I or a 109 Cd source. Measurements were taken from tibia in intact human legs, post-mortem. On the basis of preliminary measurements, it was concluded that an exposure of one rad is adequate for determination of lead in bone. Both the advantages and the disadvantages of L-x-rays, used in the technique developed for this study, are compared with those of K-x-rays

  15. Analysis of Precious Stones Deposited in Various Rock Samples of Mogok Region by energy dispersive X-ray Fluorescence Spectrometry

    International Nuclear Information System (INIS)

    Kyi Kyi San; Soe Lwin; Win Win Thar; Sein Htoon

    2004-06-01

    The analysis of precious stones deposited in various rock samples of Mogok region were investigated by the energy dispersive x-ray fluorescence technique. The x-ray machine with Rh target was used to excite the characteristic x-ray from the sample. X-rays emitted from the sample were measured by a high resolution, cooled Si (Li) detector. The calibration was made by the measurement of minerals which composed in each kind of precious stones. The kind of precious stone deposited in the rocks sample was determined by the measurement of minerals from the rock samples compared with those obtained from each kind of precious stones

  16. Non-destructive in situ study of “Mad Meg” by Pieter Bruegel the Elder using mobile X-ray fluorescence, X-ray diffraction and Raman spectrometers

    Energy Technology Data Exchange (ETDEWEB)

    Van de Voorde, Lien, E-mail: lien.vandevoorde@ugent.be [Ghent University, Department of Analytical Chemistry, X-ray Microspectroscopy and Imaging Research Group, Krijgslaan 281 S12, B-9000 Gent (Belgium); Van Pevenage, Jolien [Ghent University, Department of Analytical Chemistry, Raman Spectroscopy Research Group, Krijgslaan 281 S12, B-9000 Gent (Belgium); De Langhe, Kaat [Ghent University, Department of Archaeology, Archaeometry Research Group, Sint-Pietersnieuwstraat 35, B-9000 Gent (Belgium); De Wolf, Robin; Vekemans, Bart; Vincze, Laszlo [Ghent University, Department of Analytical Chemistry, X-ray Microspectroscopy and Imaging Research Group, Krijgslaan 281 S12, B-9000 Gent (Belgium); Vandenabeele, Peter [Ghent University, Department of Archaeology, Archaeometry Research Group, Sint-Pietersnieuwstraat 35, B-9000 Gent (Belgium); Martens, Maximiliaan P.J. [Ghent University, Department of Art, Music and Theatre Sciences, Blandijnberg 2, B-9000 Gent (Belgium)

    2014-07-01

    “Mad Meg”, a figure of Flemish folklore, is the subject of a famous oil-on-panel painting by the Flemish renaissance artist Pieter Bruegel the Elder, exhibited in the Museum Mayer van den Bergh (Antwerp, Belgium). This article reports on the in situ chemical characterization of this masterpiece by using currently available state-of-the-art portable analytical instruments. The applied non-destructive analytical approach involved the use of a) handheld X-ray fluorescence instrumentation for retrieving elemental information and b) portable X-ray fluorescence/X-ray diffraction instrumentation and laser-based Raman spectrometers for obtaining structural/molecular information. Next to material characterization of the used pigments and of the different preparation layers of the painting, also the verification of two important historical iconographic hypotheses is performed concerning the economic way of painting by Brueghel, and whether or not he used blue smalt pigment for painting the boat that appears towards the top of the painting. The pigments identified are smalt pigment (65% SiO{sub 2} + 15% K{sub 2}O + 10% CoO + 5% Al{sub 2}O{sub 3}) for the blue color present in all blue areas of the painting, probably copper resinate for the green colors, vermillion (HgS) as red pigment and lead white is used to form different colors. The comparison of blue pigments used on different areas of the painting gives no differences in the elemental fingerprint which confirms the existing hypothesis concerning the economic painting method by Bruegel. - Highlights: • In situ, non-destructive investigation of a famous painting by Pieter Bruegel. • Use of a new, commercial available, portable XRF/XRD instrumentation. • Multi-methodological approach: make also use of a mobile Raman spectrometer. • Used pigments and different preparation layers of the painting are characterized. • The verification of two important historical iconographic hypotheses are performed.

  17. Non-destructive in situ study of “Mad Meg” by Pieter Bruegel the Elder using mobile X-ray fluorescence, X-ray diffraction and Raman spectrometers

    International Nuclear Information System (INIS)

    Van de Voorde, Lien; Van Pevenage, Jolien; De Langhe, Kaat; De Wolf, Robin; Vekemans, Bart; Vincze, Laszlo; Vandenabeele, Peter; Martens, Maximiliaan P.J.

    2014-01-01

    “Mad Meg”, a figure of Flemish folklore, is the subject of a famous oil-on-panel painting by the Flemish renaissance artist Pieter Bruegel the Elder, exhibited in the Museum Mayer van den Bergh (Antwerp, Belgium). This article reports on the in situ chemical characterization of this masterpiece by using currently available state-of-the-art portable analytical instruments. The applied non-destructive analytical approach involved the use of a) handheld X-ray fluorescence instrumentation for retrieving elemental information and b) portable X-ray fluorescence/X-ray diffraction instrumentation and laser-based Raman spectrometers for obtaining structural/molecular information. Next to material characterization of the used pigments and of the different preparation layers of the painting, also the verification of two important historical iconographic hypotheses is performed concerning the economic way of painting by Brueghel, and whether or not he used blue smalt pigment for painting the boat that appears towards the top of the painting. The pigments identified are smalt pigment (65% SiO 2 + 15% K 2 O + 10% CoO + 5% Al 2 O 3 ) for the blue color present in all blue areas of the painting, probably copper resinate for the green colors, vermillion (HgS) as red pigment and lead white is used to form different colors. The comparison of blue pigments used on different areas of the painting gives no differences in the elemental fingerprint which confirms the existing hypothesis concerning the economic painting method by Bruegel. - Highlights: • In situ, non-destructive investigation of a famous painting by Pieter Bruegel. • Use of a new, commercial available, portable XRF/XRD instrumentation. • Multi-methodological approach: make also use of a mobile Raman spectrometer. • Used pigments and different preparation layers of the painting are characterized. • The verification of two important historical iconographic hypotheses are performed

  18. A note on the assay of special nuclear materials in solution by x-ray fluorescence

    International Nuclear Information System (INIS)

    Canada, T.R.; Hsue, S.T.

    1982-01-01

    Presents a formulation that allows empirical results of the ''internal standard'' approach to be understood in a quantifiable manner, and suggests an alternative measurement procedure that removes many of the technique's undesirable features while maintaining those that add to instrumental accuracy. Assumes that the reader is familiar with x-ray fluorescence (XRF) technology. Promises a more detailed presentation, including proof-of-principle experimental results, in the future. Points out that practical applications of this approach may be achieved with both K- and L-x-ray fluorescence. Concludes that the formulation and alternative measurement procedure suggested indicates that the ''internal standard'' approach may be improved by making measurements at one or more additional x-ray energies of the element to be assayed. Effects of solution acidity variations and the relative concentrations of plutonium and uranium may be avoided. Because of the inherent stability of ratio techniques, little or no modification to this formulation is anticipated for cylindrical near-field geometries

  19. Dose distribution calculation for in-vivo X-ray fluorescence scanning

    International Nuclear Information System (INIS)

    Figueroa, R. G.; Lozano, E.; Valente, M.

    2013-01-01

    In-vivo X-ray fluorescence constitutes a useful and accurate technique, worldwide established for constituent elementary distribution assessment. Actually, concentration distributions of arbitrary user-selected elements can be achieved along sample surface with the aim of identifying and simultaneously quantifying every constituent element. The method is based on the use of a collimated X-ray beam reaching the sample. However, one common drawback for considering the application of this technique for routine clinical examinations was the lack of information about associated dose delivery. This work presents a complete study of the dose distribution resulting from an in-vivo X-ray fluorescence scanning for quantifying biohazard materials on human hands. Absorbed dose has been estimated by means of dosimetric models specifically developed to this aim. In addition, complete dose distributions have been obtained by means of full radiation transport calculations in based on stochastic Monte Carlo techniques. A dedicated subroutine has been developed using the Penelope 2008 main code also integrated with dedicated programs -Mat Lab supported- for 3 dimensional dose distribution visualization. The obtained results show very good agreement between approximate analytical models and full descriptions by means of Monte Carlo simulations. (Author)

  20. Determination of minor and trace elements in kidney stones by x-ray fluorescence analysis

    Science.gov (United States)

    Srivastava, Anjali; Heisinger, Brianne J.; Sinha, Vaibhav; Lee, Hyong-Koo; Liu, Xin; Qu, Mingliang; Duan, Xinhui; Leng, Shuai; McCollough, Cynthia H.

    2014-03-01

    The determination of accurate material composition of a kidney stone is crucial for understanding the formation of the kidney stone as well as for preventive therapeutic strategies. Radiations probing instrumental activation analysis techniques are excellent tools for identification of involved materials present in the kidney stone. In particular, x-ray fluorescence (XRF) can be very useful for the determination of minor and trace materials in the kidney stone. The X-ray fluorescence measurements were performed at the Radiation Measurements and Spectroscopy Laboratory (RMSL) of department of nuclear engineering of Missouri University of Science and Technology and different kidney stones were acquired from the Mayo Clinic, Rochester, Minnesota. Presently, experimental studies in conjunction with analytical techniques were used to determine the exact composition of the kidney stone. A new type of experimental set-up was developed and utilized for XRF analysis of the kidney stone. The correlation of applied radiation source intensity, emission of X-ray spectrum from involving elements and absorption coefficient characteristics were analyzed. To verify the experimental results with analytical calculation, several sets of kidney stones were analyzed using XRF technique. The elements which were identified from this techniques are Silver (Ag), Arsenic (As), Bromine (Br), Chromium (Cr), Copper (Cu), Gallium (Ga), Germanium (Ge), Molybdenum (Mo), Niobium (Nb), Rubidium (Rb), Selenium (Se), Strontium (Sr), Yttrium (Y), Zirconium (Zr). This paper presents a new approach for exact detection of accurate material composition of kidney stone materials using XRF instrumental activation analysis technique.

  1. Fluorescent X-ray computed tomography using synchrotron radiation for imaging nonradioactive tracer materials

    Energy Technology Data Exchange (ETDEWEB)

    Akiba, Masahiro; Yuasa, Tetsuya; Uchida, Akira; Akatsuka, Takao [Yamagata Univ., Yonezawa (Japan). Electrical and Information of Engineering; Takeda, Tohoru; Hyodo, Kazuyuki; Itai, Yuji

    1997-09-01

    We describe a system of fluorescent X-ray computed tomography using synchrotron radiation (SR-FXCT) to image nonradioactive contrast materials. The system operates on the basis of computed tomography (CT) scanned by the pencil beam. In the previous experiment, we have imaged an acrylic cylindrical phantom with cross-shaped channel, filled with a diluted iodine-based tracer material of 200 {mu}g/ml. This research is aimed to improve image quality, to select the optimum energy of the incident X-ray, to confirm quantitative evaluation of the image, and to demonstrate FXCT image for living body. First, we simulated output energy profile by the Monte Carlo simulation and confirmed to predetermine the incident X-ray energy at 37 keV, in order to separate the fluorescent photons from background scattering components. Next, the imaging experiment was performed by using conventional CT algorithm under the optimum parameter at the Tristan Accumulation Ring, KEK, Japan. An acrylic phantom containing five paraxial channels of 5 and 4 mm in diameter, could be imaged; where each channel was respectively filled with diluted iodine-based contrast materials of 50, 100, 200 and 500 {mu}g/ml. From the reconstructed image, we confirmed quantitativity in the FXCT image. Finally, a rat`s brain was imaged in vitro by FXCT and monochromatic transmission CT. The comparison between these results showed that the iodine-rich region in the FXCT image corresponded with that in the monochromatic transmission CT image. (author)

  2. X-ray fluorescence spectroscopy technology applied to the materials elementary characterization

    International Nuclear Information System (INIS)

    Marambio A, Cristian Gilberto.

    1997-01-01

    A thorough study of the different applications of energy dispersive x-ray fluorescence spectrometry is presented, using different excitation sources and measurement geometries. The adaptation of these systems focuses on the analytical solution for different sample types by studying distinct parameters such as: volume of saturation for liquid samples, inter elemental effects from the matrix and the interferences associated with the measurement reading statistical parameters: as reproducibility, precision and detection limits. The application of the technique using radioisotopic sources gave satisfactory results in the analysis of geologic samples, in analytical control of concentration processes for rare earths and for the manufacturing of fuel elements. In the case of a system with an x-ray generator two measurement geometries were studied: 45 deg geometry and total reflection. There were major results in the analysis of polymer impurities and alloy impurities, aluminums and thin semiconductor films, respectively, after non destructive in situ analysis of the material. The results show that x-ray fluorescence spectrometry is a powerful tool for analysis and process control, with prospects for the solution of analytical problems in the materials area. (author)

  3. Measurement of plutonium in spent nuclear fuel by self-induced x-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Hoover, Andrew S [Los Alamos National Laboratory; Rudy, Cliff R [Los Alamos National Laboratory; Tobin, Steve J [Los Alamos National Laboratory; Charlton, William S [Los Alamos National Laboratory; Stafford, A [TEXAS A& M; Strohmeyer, D [TEXAS A& M; Saavadra, S [ORNL

    2009-01-01

    Direct measurement of the plutonium content in spent nuclear fuel is a challenging problem in non-destructive assay. The very high gamma-ray flux from fission product isotopes overwhelms the weaker gamma-ray emissions from plutonium and uranium, making passive gamma-ray measurements impossible. However, the intense fission product radiation is effective at exciting plutonium and uranium atoms, resulting in subsequent fluorescence X-ray emission. K-shell X-rays in the 100 keV energy range can escape the fuel and cladding, providing a direct signal from uranium and plutonium that can be measured with a standard germanium detector. The measured plutonium to uranium elemental ratio can be used to compute the plutonium content of the fuel. The technique can potentially provide a passive, non-destructive assay tool for determining plutonium content in spent fuel. In this paper, we discuss recent non-destructive measurements of plutonium X-ray fluorescence (XRF) signatures from pressurized water reactor spent fuel rods. We also discuss how emerging new technologies, like very high energy resolution microcalorimeter detectors, might be applied to XRF measurements.

  4. Application of X-ray fluorescence (WDXRF): thickness and chemical composition determination of thin films

    International Nuclear Information System (INIS)

    Scapin, Valdirene de Oliveira.

    2004-01-01

    In this work a procedure is described for thickness and quantitative chemical composition of thin films by wavelength dispersion X-ray fluorescence (WDXRF) using Fundamental Parameters method. This method was validated according to quality assurance standard and applied sample Al, Cr, TiO2, Ni, ZrO2 (single thickness) and Ni/Cr (double thickness) on glass; Ni on steel and metallic zinc and TiO2 on metallic iron (single thickness), all the sample were prepared for physical deposition of vapor (PVD). The thickness had been compared with Absorption (FRX-A) and Rutherford Backscattering Spectrometry (RBS) methods; the result showed good efficiency of the fundamental parameters method. Sample structural characteristics analyzed by X ray diffraction (XRD) showed any influence in the thickness determinations. (author)

  5. The studies of post-medieval glass by multivariate and X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Kierzek, J.; Kunicki-Goldfinger, J.

    2002-01-01

    Multivariate statistical analysis of the results obtained by energy dispersive X-ray fluorescence analysis has been used in the study of baroque vessel glasses originated from central Europe. X-ray spectrometry can be applied as a completely non-destructive, non-sampling and multi-element method. It is very useful in the studies of valuable historical artefacts. For the last years, multivariate statistical analysis has been developed as an important tool for the archaeometric purposes. Cluster, principal component and discriminant analysis were applied for the classification of the examined objects. The obtained results show that these statistical tools are very useful and complementary in the studies of historical objects. (author)

  6. Metals determination in wood treated by synchrotron radiation X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Vives, Ana Elisa Sirito de; Medeiros, Jean Gabriel da Silva; Tomazello Filho, Mario

    2005-01-01

    The paper describes the use of X-Ray fluorescence analysis for distribution and quantification of metals in the hardwood (Eucalyptus sp) and softwood (Pinus sp) treated with CCA (copper-chromium-arsenic). The sapwood/heartwood for hardwood sample and the growth-rings for softwood sample were analyzed. The samples were scanned in 320 mm steps in the vertical direction. For excitation of the elements a white beam synchrotron radiation of ∼ 320 x 180 mm was employed and for the X-ray detection a Si(Li) semiconductor detector. The elements K, Ca, Cr, Mn, Cu, Zn and As were determined. Fundamental parameters were used to quantify the elements concentrations. (author)

  7. Thin coating thickness determination using radioisotope-excited x-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Del Castillo, Lorena A.; Calix, Virginia S.

    2001-01-01

    Three different approaches on thin coating thickness determination using a radioisotope-excited x-ray fluorescence spectrometry were demonstrated and results were compared. A standard of thin layer of gold (Au) on a nickel (Ni) substrate from the US National Bureau of Standards (with a nominal thickness of 0.300505 microns of at least 99.9% Au electrodeposited over 2 nils of Ni) on low carbon steel (1010) was analyzed using a Cd 109-excited XRF system. Au thickness computations were done using the (a) thin standard approach, (b) thick standard approach, and (c) x-ray absorption method (ASTM A754-79 1982). These three methods yielded results within the limit set by the American Society for Testing Materials (ASTM), which is +/-3%. Of the three methods, the thick standard yielded the best result with 0.124% error. (Author)

  8. Physical aspects of quantitative particles analysis by X-ray fluorescence and electron microprobe techniques

    International Nuclear Information System (INIS)

    Markowicz, A.

    1986-01-01

    The aim of this work is to present both physical fundamentals and recent advances in quantitative particles analysis by X-ray fluorescence (XRF) and electron microprobe (EPXMA) techniques. A method of correction for the particle-size effect in XRF analysis is described and theoretically evaluated. New atomic number- and absorption correction procedures in EPXMA of individual particles are proposed. The applicability of these two correction methods is evaluated for a wide range of elemental composition, X-ray energy and sample thickness. Also, a theoretical model for composition and thickness dependence of Bremsstrahlung background generated in multielement bulk specimens as well as thin films and particles are presented and experimantally evaluated. Finally, the limitations and further possible improvements in quantitative particles analysis by XFR and EPXMA are discussed. 109 refs. (author)

  9. Determination of low concentrations of thorium in granites using X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Shigematsu, H.M.; Sato, I.M.; Iyer, S.S.

    1981-03-01

    An analytical method for the accurate determination of low concentrations of thorium in rocks using X-ray fluorescence technique, was developed. A tungsten tube was utilized for the production of X-rays. The samples were prepared in the form of double layer pressed pellets using boric acid as a binding agent. The concentration of thorium was determined by measuring the intensity of the characteristic first order Th L α line. The calibration was carried out with USGS rock standards AGV-1, GSP-1 and G-2. Seven granite rocks samples from Granite Mountains of Wyoming, USA, supplied by Dr. Stuckless. Also were analysed. The results obtained were compared with values obtained in others laboratories using different analytical methods. The analyses show that the thorium is concentrated in accessory minerals and presented a non-uniform distribution, making sampling an important factor in the analysis of thorium. A discussion of the precision and accuracy of the method is presented. (Author) [pt

  10. Elemental investigation on Spanish dinosaur bones by x-ray fluorescence

    International Nuclear Information System (INIS)

    Brunetti, Antonio; Golosio, Bruno; Stegel, Giovanni; Piga, Giampaolo; Lasio, Barbara; Oliva, Piernicola; Enzo, Stefano

    2013-01-01

    In this paper we examine the chemical composition results obtained on a collection of 18 dinosaur fossil bones from Spain studied using a portable x-ray fluorescence spectrometer together with a reverse Monte Carlo numerical technique of data analysis. This approach is applied to the hypothesis of arbitrarily rough surfaces in order to account for the influence of the surface state of specimens on the chemical content evaluation. It is confirmed that the chemical content of elements is essential for understanding the changes brought about by diagenetic and taphonomic processes. However, for precise knowledge of what changes fossil bones have undergone after animal life and burial, it is necessary to use a multi-technique approach making use of other instruments like x-ray diffraction in order to describe accurately the transformations undergone by the mineralogical and bioinorganic phases and the properties of specific molecular groups. (paper)

  11. A wavelet-based Gaussian method for energy dispersive X-ray fluorescence spectrum

    Directory of Open Access Journals (Sweden)

    Pan Liu

    2017-05-01

    Full Text Available This paper presents a wavelet-based Gaussian method (WGM for the peak intensity estimation of energy dispersive X-ray fluorescence (EDXRF. The relationship between the parameters of Gaussian curve and the wavelet coefficients of Gaussian peak point is firstly established based on the Mexican hat wavelet. It is found that the Gaussian parameters can be accurately calculated by any two wavelet coefficients at the peak point which has to be known. This fact leads to a local Gaussian estimation method for spectral peaks, which estimates the Gaussian parameters based on the detail wavelet coefficients of Gaussian peak point. The proposed method is tested via simulated and measured spectra from an energy X-ray spectrometer, and compared with some existing methods. The results prove that the proposed method can directly estimate the peak intensity of EDXRF free from the background information, and also effectively distinguish overlap peaks in EDXRF spectrum.

  12. Metals determination in wood treated by synchrotron radiation X-ray fluorescence analysis

    Energy Technology Data Exchange (ETDEWEB)

    Vives, Ana Elisa Sirito de [Universidade Metodista de Piracicaba (UNIMEP), Santa Barbara D' Oeste, SP (Brazil). Faculdade de Engenharia, Arquitetura e Urbanismo]. E-mail: aesvives@unimep.br; Silva, Richard Maximiliano da Cunha [Centro de Energia Nuclear na Agricultura, Piracicaba, SP (Brazil)]. E-mail: maxcunha@cena.usp.br; Medeiros, Jean Gabriel da Silva; Tomazello Filho, Mario [Sao Paulo Univ., Piracicaba, SP (Brazil). Escola Superior de Agricultura Luiz de Queiroz]. E-mail: jeangm@esalq.usp.br; mtomazel@esalq.usp.br; Moreira, Silvana [Universidade Estadual de Campinas, SP (Brazil). Faculdade de Engenharia Civil, Arquitetura e Urbanismo]. E-mail: Silvana@fec.unicamp.br; Zucchi, Orgheda Luiza Araujo Domingues [Sao Paulo Univ., Ribeirao Preto, SP (Brazil). Faculdade de Ciencias Farmaceuticas]. E-mail: olzucchi@fcfrp.usp.br; Barroso, Regina Cely [Universidade do Estado, Rio de Janeiro, RJ (Brazil)]. E-mail: cely@uerj.br

    2005-07-01

    The paper describes the use of X-Ray fluorescence analysis for distribution and quantification of metals in the hardwood (Eucalyptus sp) and softwood (Pinus sp) treated with CCA (copper-chromium-arsenic). The sapwood/heartwood for hardwood sample and the growth-rings for softwood sample were analyzed. The samples were scanned in 320 mm steps in the vertical direction. For excitation of the elements a white beam synchrotron radiation of {approx} 320 x 180 mm was employed and for the X-ray detection a Si(Li) semiconductor detector. The elements K, Ca, Cr, Mn, Cu, Zn and As were determined. Fundamental parameters were used to quantify the elements concentrations. (author)

  13. Radiography imaging of cultural heritage obtained with a modified portable X-Ray Fluorescence System

    International Nuclear Information System (INIS)

    Mendoza Cuevas, Ariadna; Velazquez Maldonado, Luis Ramon

    2011-01-01

    The sufficiency of imaging quality of the radiographies obtained with a modified portable X-ray fluorescence spectrometer was evaluated for the study of cultural heritage. The proposed instrument use an X-ray tube with Pd anode (2 mm) that allows a maximum voltage and current of 50 kV and 1 mA respectively and a collimation system permit to irradiate a square shape region in the analyzed object by the projection of light beam with the same shape on its surface. The spatial resolution of the obtained radiographic image make possible to localize and well define pentimenti in painting, identify filling materials in a painting under restoration process, the radiogrametry of archaeological bone and the identification of a petrified sphere from an archaeological discovery. The radiographic analysis is proposed for study of physical anthropology in Cuba. (author)

  14. Reverse engineering the ancient ceramic technology based on X-ray fluorescence spectromicroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Sciau, Philippe; Leon, Yoanna; Goudeau, Philippe; Fakra, Sirine C.; Webb, Sam; Mehta, Apurva

    2011-07-06

    We present results of X-ray fluorescence (XRF) microprobe analyses of ancient ceramic cross-sections aiming at deciphering the different firing protocols used for their production. Micro-focused XRF elemental mapping, Fe chemical mapping and Fe K-edge X-ray absorption near edge structure spectroscopy were performed on pre-sigillata ceramics from southern Gaul, and terra Sigillata vessels from Italy and southern Gaul. Pieces from the different workshops and regions showed significant difference in the starting clay material, clay conditioning and kiln firing condition. By contrast, sherds from the same workshop exhibited more subtle differences and possible misfirings. Understanding the precise firing conditions and protocols would allow recreation of kilns for various productions. Furthermore, evolution and modification of kiln design would shed some light on how ancient potters devised solutions to diverse technological problems they encountered.

  15. Quantitative X-ray fluorescence analysis at the ESRF ID18F microprobe

    CERN Document Server

    Vekemans, B; Somogyi, A; Drakopoulos, M; Kempenaers, L; Simionovici, A; Adams, F

    2003-01-01

    The new ID18F end-station at the European synchrotron radiation facility (ESRF) in Grenoble (France) is dedicated to sensitive and accurate quantitative micro-X-ray fluorescence (XRF) analysis at the ppm level with accuracy better than 10% for elements with atomic numbers above 18. For accurate quantitative analysis, given a high level of instrumental stability, major steps are the extraction and conversion of experimental X-ray line intensities into elemental concentrations. For this purpose a two-step quantification approach was adopted. In the first step, the collected XRF spectra are deconvoluted on the basis of a non-linear least-squares fitting algorithm (AXIL). The extracted characteristic line intensities are then used as input for a detailed Monte Carlo (MC) simulation code dedicated to XRF spectroscopy taking into account specific experimental conditions (excitation/detection) as well as sample characteristics (absorption and enhancement effects, sample topology, heterogeneity etc.). The iterative u...

  16. Development of a reconstruction software of elemental maps by micro X-ray fluorescence

    International Nuclear Information System (INIS)

    Almeida, Andre Pereira de; Braz, Delson; Mota, Carla Lemos; Oliveira, Luis Fernando de; Barroso, Regina Cely; Pinto, Nivia Graciele Villela; Cardoso, Simone Coutinho; Moreira, Silvana

    2009-01-01

    The technique of X-ray fluorescence (XRF) using SR microbeams is a powerful analysis tool for studying elemental composition in several samples. One application of this technique is the analysis done through the mapping of chemical elements forming a matrix of data. The aim of this work is the presentation of the program MapXRF, an in-house software designed to optimize the processing and mapping of fluorescence intensities data. This program uses spectra generated by QXAS as input data and separates the intensities of each chemical element found in the fluorescence spectra in files themselves. From these files, the program generates the intensity maps that can be visualized in any program of treatment of images. The proposed software was tested using fluorescence data obtained in the XRF beamline at National Synchrotron Light Laboratory (LNLS), Brazil. Automatic 2D scans were performed and element distribution maps were obtained in form of a matrix of data. (author)

  17. Development of a software for reconstruction of X-ray fluorescence intensity maps

    International Nuclear Information System (INIS)

    Almeida, Andre Pereira de; Braz, Delson; Mota, Carla Lemos; Oliveira, Luis Fernando de; Barroso, Regina Cely; Pinto, Nivia Graciele Villela; Cardoso, Simone Coutinho; Moreira, Silvana

    2009-01-01

    The technique of X-ray fluorescence (XRF) using SR microbeams is a powerful analysis tool for studying elemental composition in several samples. One application of this technique is the analysis done through the mapping of chemical elements forming a matrix of data. The aim of this work is the presentation of the program MapXRF, an in-house software designed to optimize the processing and mapping of fluorescence intensities data. This program uses spectra generated by QXAS as input data and separates the intensities of each chemical element found in the fluorescence spectra in files themselves. From these files, the program generates the intensity maps that can be visualized in any program of treatment of images. The proposed software was tested using fluorescence data obtained in the XRF beamline of XRF at Synchrotron Light National Laboratory (LNLS), Brazil. Automatic 2D scans were performed and element distribution maps were obtained in the form of a matrix of data. (author)

  18. Development of a software for reconstruction of X-ray fluorescence intensity maps

    Energy Technology Data Exchange (ETDEWEB)

    Almeida, Andre Pereira de; Braz, Delson; Mota, Carla Lemos, E-mail: apalmeid@gmail.co, E-mail: delson@lin.ufrj.b, E-mail: clemos@con.ufrj.b [Coordenacao dos Programas de Pos-graduacao de Engenharia (COPPE/UFRJ), Rio de Janeiro, RJ (Brazil). Programa de Engenharia Nuclear; Oliveira, Luis Fernando de; Barroso, Regina Cely; Pinto, Nivia Graciele Villela, E-mail: cely@uerj.b, E-mail: lfolive@uerj.b, E-mail: nitatag@gmail.co [Universidade do Estado do Rio de Janeiro (UERJ), Rio de Janeiro, RJ (Brazil). Inst. de Fisica; Cardoso, Simone Coutinho, E-mail: simone@if.ufrj.b [Universidade Federal do Rio de Janeiro (UFRJ), RJ (Brazil). Inst. de Fisica; Moreira, Silvana, E-mail: silvana@fec.unicamp.b [Universidade Estadual de Campinas (UNICAMP), SP (Brazil). Faculdade de Engenharia Civil, Arquitetura e Urbanismo

    2009-07-01

    The technique of X-ray fluorescence (XRF) using SR microbeams is a powerful analysis tool for studying elemental composition in several samples. One application of this technique is the analysis done through the mapping of chemical elements forming a matrix of data. The aim of this work is the presentation of the program MapXRF, an in-house software designed to optimize the processing and mapping of fluorescence intensities data. This program uses spectra generated by QXAS as input data and separates the intensities of each chemical element found in the fluorescence spectra in files themselves. From these files, the program generates the intensity maps that can be visualized in any program of treatment of images. The proposed software was tested using fluorescence data obtained in the XRF beamline of XRF at Synchrotron Light National Laboratory (LNLS), Brazil. Automatic 2D scans were performed and element distribution maps were obtained in the form of a matrix of data. (author)

  19. Development of a reconstruction software of elemental maps by micro X-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Almeida, Andre Pereira de; Braz, Delson; Mota, Carla Lemos, E-mail: apalmeid@gmail.co, E-mail: delson@lin.ufrj.b, E-mail: clemos@con.ufrj.b [Universidade Federal do Rio de Janeiro (PEN/COPPE/UFRJ), RJ (Brazil). Coordenacao dos Programas de Pos-Graduacao de Engenharia. Programa de Energia Nuclear; Oliveira, Luis Fernando de; Barroso, Regina Cely; Pinto, Nivia Graciele Villela, E-mail: cely@uerj.b, E-mail: lfolive@uerj.b, E-mail: nitatag@gmail.co [Universidade do Estado do Rio de Janeiro (IF/UERJ), RJ (Brazil). Inst. de Fisica; Cardoso, Simone Coutinho [Universidade Federal do Rio de Janeiro (IF/UFRJ), RJ (Brazil). Inst. de Fisica; Moreira, Silvana [Universidade Estadual de Campinas (FEC/UNICAMP), SP (Brazil) Faculdade de Engenharia Civil, Arquitetura e Urbanismo

    2009-07-01

    The technique of X-ray fluorescence (XRF) using SR microbeams is a powerful analysis tool for studying elemental composition in several samples. One application of this technique is the analysis done through the mapping of chemical elements forming a matrix of data. The aim of this work is the presentation of the program MapXRF, an in-house software designed to optimize the processing and mapping of fluorescence intensities data. This program uses spectra generated by QXAS as input data and separates the intensities of each chemical element found in the fluorescence spectra in files themselves. From these files, the program generates the intensity maps that can be visualized in any program of treatment of images. The proposed software was tested using fluorescence data obtained in the XRF beamline at National Synchrotron Light Laboratory (LNLS), Brazil. Automatic 2D scans were performed and element distribution maps were obtained in form of a matrix of data. (author)

  20. Study of uranium contamination of ground water in Punjab using X-ray fluorescence technique

    Energy Technology Data Exchange (ETDEWEB)

    Alrakabi, Muhanad; Singh, Gurjeet; Bhalla, Atul; Kumar, Sunil; Kumar, Sanjeev; Rai, Bimal; Singh, N; Shahi, J S; Mehta, D [Department of Physics, Panjab University, Chandigarh (India); Srivastava, Alok [Department of Chemistry, Panjab University, Chandigarh (India)

    2010-07-01

    A number of reports have appeared in public media about uranium ingestion being a possible cause for cancer and increased birth rate abnormalities among children in the Malwa region of Punjab state in India. These reports link problems like cancer and Autism, with the presence of uranium in the ground waters of Malwa region. The concentration of uranium in drinking water from sources as varied as ground water, canal water supply and reverse osmosis system have been investigated using X-ray fluorescence technique. Samples from the thermal power plants in the regions and nearby ground waters were also analyzed to identify the source of contamination. The samples were collected with assistance of the officials from the Government of Punjab. More than half a litre of each of the water samples was dried at 60 deg-80 deg in an oven. Residue was collected using larger quantities of water samples in case of RO water samples. The elemental analysis of the residue was carried out using the Energy-Dispersive X-Ray Fluorescence (EDXRF) spectrometer consisting of an {sup 42}Mo-anode X-ray tube (Panalytical, 2.5 kW) as an excitation source and a Si(Li) detector. A combination of selective absorbers of {sup 30}Zn, {sup 38}Sr, and {sup 39}Y was used in the incident beam for improving the detection limit for Uranium by reducing the background and removing the {sup 42}Mo K X-rays. The detection limit in ppb/litre depends upon the amount of residue

  1. Anomalous scattering, transport, and spatial distribution of X-ray fluorescence at the exit of polycapillary structures

    Energy Technology Data Exchange (ETDEWEB)

    Mazuritskiy, M. I., E-mail: mazurmik@gmail.com; Lerer, A. M.; Makhno, P. V. [Southern Federal University (Russian Federation)

    2016-12-15

    The angular distribution of the X-ray intensity at the exit of microchannel plates at grazing incidence of monochromatic radiation on the walls of microcapillaries has been investigated. The angles and energies of the primary radiation quanta at which the synchrotron beam excites X-ray fluorescence propagating inside polycapillary structures have been determined. The angular dependences of the intensity distribution of X-rays transmitted through the microcapillaries have been studied theoretically and experimentally for energies corresponding to the region of anomalous dispersion near the L{sub 2,3} absorption edges of silicon. The propagation of waves in hollow polycapillary waveguides, the excitation of X-ray fluorescence, and the X-ray diffraction at the exit of microchannel plates have been modeled mathematically. The mathematical model takes into account the presence of a transition layer on the microchannel surface.

  2. Use of fluorescent-metal intensifying screens with RT-type films for X-ray radiography using pulse devices

    International Nuclear Information System (INIS)

    Morgovskij, L.Ya.; Khakim'yanov, R.R.

    1985-01-01

    A study was made on characteristics of combination of fluorescent-metal Kyokko SMP-308 (Japan) and RCF (Agfa-Gevert) screens with domestic X-ray RT-1, RT-2, RT-5 films. Pulse X-ray MIRA-3D and NORA devices at 200 kV voltage amplitude in X-ray tube were used as radiation source. Testing was conducted for steel samples of 5-40 mm thickness. Comparative exposures for various film combinations with fluorescent-metal screens, fluorescent VP-2 screens and lead foils of 27 μm thickness were determined at that. It is shown that fluorescent-metal screens can be successfully applied with domestic X-ray technical films. They enable to decrease exposure by one order with insignificant deterioration of sensitivity. It is important for testing of pipeline welds

  3. Simultaneous analyses of gaseous and particulate sulphur compounds in the atmosphere by x-ray fluorescence spectrometry, (1)

    International Nuclear Information System (INIS)

    Matsuda, Yatsuka; Mamuro, Tetsuo

    1974-01-01

    An analytical technique for the simultaneous measurements of the atmospheric concentrations of SO 2 gas and sulphur absorbed by aerosol particles has been developed. Aerosol particles are collected on membrane filter and at the same time SO 2 gas is captured on alkali impregnated filter. The sulphur content in each filter is measured by an energy dispersive X-ray fluorescence spectrometer consisting of a Si(Li) semiconductor detector connected to a multichannel pulse hight analyzer and an excitation source of 55 Fe. Two methods are acceptable for the determination of the sulphur content in impregnated filter by X-ray fluorescence analysis. In the first method X-ray fluorescence analysis is made after the collected sulphur has diffused and distributed uniformly enough throughout filter, and in the second method X-ray fluorescence analysis has to be finished before the diffusion of the collected sulphur becomes appreciable. (auth.)

  4. Study of the elemental composition of Chenopodium Quinoa Willd by fast neutron activation analysis and X ray fluorescence analysis

    International Nuclear Information System (INIS)

    Soto Moran, R.L.; Szegedi, S.; Llopiz, J.L.

    1996-01-01

    By means of x-ray fluorescence and fast neutron activation analysis the nitrogen content has been determined in samples of roots, stems, leaf, flowers and grains from Quinua (Chenopodium Quinoa Willd), which was previously treated with fertilizer

  5. Development of glancing-incidence and glancing-take-off X-ray fluorescence apparatus for surface and thin-film analyses

    International Nuclear Information System (INIS)

    Tsuji, Kouichi; Wagatsuma, Kazuaki; Yamada, Takashi; Utaka, Tadashi

    1997-01-01

    We have studied X-ray fluorescence analysis under glancing incidence and glancing take-off conditions. Recently, we have developed a third apparatus for detecting glancing-incidence and take-off X-ray fluorescence, which makes it possible to measure the incident-angle dependence, the take-off-angle dependence. X-ray reflectivity, and X-ray diffraction. Primarily, we have measured the take-off angular dependence of X-ray fluorescence using this apparatus. Glancing take-off X-ray fluorescence has some advantages in comparison with glancing-incidence X-ray fluorescence. The surface density and the absolute angles were determined by analysing the take-off angle dependence of the fluorescent X-rays emitted from identical atoms with the aid of the reciprocity theorem. (Author)

  6. Incoherent-scatter computed tomography with monochromatic synchrotron x ray: feasibility of multi-CT imaging system for simultaneous measurement-of fluorescent and incoherent scatter x rays

    Science.gov (United States)

    Yuasa, T.; Akiba, M.; Takeda, T.; Kazama, M.; Hoshino, A.; Watanabe, Y.; Hyodo, K.; Dilmanian, F. A.; Akatsuka, T.; Itai, Y.

    1997-10-01

    We describe a new system of incoherent scatter computed tomography (ISCT) using monochromatic synchrotron X rays, and we discuss its potential to be used in in vivo imaging for medical use. The system operates on the basis of computed tomography (CT) of the first generation. The reconstruction method for ISCT uses the least squares method with singular value decomposition. The research was carried out at the BLNE-5A bending magnet beam line of the Tristan Accumulation Ring in KEK, Japan. An acrylic cylindrical phantom of 20-mm diameter containing a cross-shaped channel was imaged. The channel was filled with a diluted iodine solution with a concentration of 200 /spl mu/gI/ml. Spectra obtained with the system's high purity germanium (HPGe) detector separated the incoherent X-ray line from the other notable peaks, i.e., the iK/sub /spl alpha// and K/sub /spl beta/1/ X-ray fluorescent lines and the coherent scattering peak. CT images were reconstructed from projections generated by integrating the counts In the energy window centering around the incoherent scattering peak and whose width was approximately 2 keV. The reconstruction routine employed an X-ray attenuation correction algorithm. The resulting image showed more homogeneity than one without the attenuation correction.

  7. Parameter study of self-absorption effects in Total Reflection X-ray Fluorescence-X-ray Absorption Near Edge Structure analysis of arsenic

    International Nuclear Information System (INIS)

    Meirer, F.; Pepponi, G.; Streli, C.; Wobrauschek, P.; Kregsamer, P.; Zoeger, N.; Falkenberg, G.

    2008-01-01

    Total reflection X-ray Fluorescence (TXRF) analysis in combination with X-ray Absorption Near Edge Structure (XANES) analysis is a powerful method to perform chemical speciation studies at trace element levels. However, when measuring samples with higher concentrations and in particular standards, damping of the oscillations is observed. In this study the influence of self-absorption effects on TXRF-XANES measurements was investigated by comparing measurements with theoretical calculations. As(V) standard solutions were prepared at various concentrations and dried on flat substrates. The measurements showed a correlation between the damping of the oscillations and the As mass deposited. A Monte-Carlo simulation was developed using data of the samples shapes obtained from confocal white light microscopy. The results showed good agreement with the measurements; they confirmed that the key parameters are the density of the investigated atom in the dried residues and the shape of the residue, parameters that combined define the total mass crossed by a certain portion of the incident beam. The study presents a simple approach for an a priori evaluation of the self-absorption in TXRF X-ray absorption studies. The consequences for Extended X-ray Absorption Fine Structure (EXAFS) and XANES measurements under grazing incidence conditions are discussed, leading to the conclusion that the damping of the oscillations seems to make EXAFS of concentrated samples non feasible. For XANES 'fingerprint' analysis samples should be prepared with a deposited mass and sample shape leading to an acceptable absorption for the actual investigation

  8. A structural study of bone changes in knee osteoarthritis by synchrotron-based X-ray fluorescence and X-ray absorption spectroscopy techniques

    Science.gov (United States)

    Sindhupakorn, Bura; Thienpratharn, Suwittaya; Kidkhunthod, Pinit

    2017-10-01

    Osteoarthritis (OA) is characterized by degeneration of articular cartilage and thickening of subchondral bone. The present study investigated the changing of biochemical components of cartilage and bone compared between normal and OA people. Using Synchrotron-based X-ray fluorescence (SR-XRF) and X-ray absorption spectroscopy (XAS) techniquesincluding X-ray absorption near edge structure (XANES) and extended X-ray absorption fine structure (EXAFS) were employed for the bone changes in kneeosteoarthritisstudies. The bone samples were collected from various osteoarthritis patients with both male and female in the ages range between 20 and 74 years old. SR-XRF results excited at 4240 eV for Ca elements show a majority three main groups, based on their XRF intensities, 20-36 years, 40-60 years and over 70 years, respectively. By employing XAS techniques, XANES features can be used to clearly explain in term of electronic transitions occurring in bone samples which are affected from osteoarthritis symptoms. Moreover, a structural change around Ca ions in bone samples is obviously obtained by EXAFS results indicating an increase of Ca-amorphous phase when the ages increase.

  9. Quantitative comparison of X-ray fluorescence microtomography setups: Standard and confocal collimator apparatus

    Energy Technology Data Exchange (ETDEWEB)

    Chukalina, M. [Institute of Microelectronics Technology RAS, 142432, Chernogolovka, Moscow District (Russian Federation)], E-mail: marina@ipmt-hpm.ac.ru; Simionovici, A. [Laboratoire de Geophysique Interne et Tectonophysique, University of Grenoble, BP 53, 38041, Grenoble (France)], E-mail: alexandre.simionovici@ujf-grenoble.fr; Zaitsev, S. [Institute of Microelectronics Technology RAS, 142432, Chernogolovka, Moscow District (Russian Federation)], E-mail: zaitsev@ipmt-hpm.ac.ru; Vanegas, C.J. [Institute of Microelectronics Technology RAS, 142432, Chernogolovka, Moscow District (Russian Federation)], E-mail: vanegas@ipmt-hpm.ac.ru

    2007-07-15

    Recently, there has been a renewed interest for fluorescence spectroscopy, as provided by modern setups which allow 2D and 3D imaging of elemental distributions. Two directions are currently under development: the SR-based fluorescence tomography in polar scanning geometry, provided by the new generation of X-ray microprobes and the confocal scanning geometry, which can be fielded in both SR and laboratory environments. The new probes bring forth a new age in fluorescence spectrometry: high resolution, high intensity and high sensitivity which allow 3D elemental mapping of volumes. The major task now is the development of these complex tools into fully quantitative probes, reproducible and straightforward for general use. In this work we analyze two X-ray fluorescence microtomography techniques: an apparatus tomography using a confocal collimator for the data collection and a standard first generation Computed Tomography (CT) in the parallel scanning scheme. We calculate the deposited dose (amount of energy deposited and distributed in the sample during the data collection time) and find the conditions for the choice of the tomography scheme.

  10. Performance comparison of two Olympus InnovX handheld x-ray analyzers for feasibility of measuring arsenic in skin in vivo – Alpha and Delta models

    International Nuclear Information System (INIS)

    Desouza, E.D.; Gherase, M.R.; Fleming, D.E.B.; Chettle, D.R.; O’Meara, J.M.; McNeill, F.E.

    2017-01-01

    The Figure-Of-Merit (FOM) performance, a combination of detection limit and dose, is compared between two generations of handheld X-Ray Fluorescence (XRF) spectrometers for the feasibility of in vivo XRF measurement of arsenic (As) in skin. The Olympus InnovX Delta model analyzer (40 kVp using either 37 or 17 μA) was found to show improvements in Minimum Detection Limit (MDL) using arsenic As-doped skin calibration phantoms with bulk tissue backing, when compared to the first generation InnovX Alpha model (40 kVp, 20 μA) in 120 s measurements. Differences between two different definitions of MDL are discussed. On the Delta system, an MDL of (0.462±0.002) μg/g As was found in phantoms, with a nylon backing behind to mimic bulk tissue behind skin. The equivalent and effective doses were found to be (10±2) mSv and ~7×10 −3 μSv respectively for the Alpha and (15±4) mSv and ~8×10 −3 μSv respectively for the Delta system in 120 s exposures. Combining MDL and effective dose, a lower (better) FOM was found for the Delta, (1.7±0.4) ppm mSv 1/2 , compared to (4.4±0.5) ppm mSv 1/2 for the Alpha model system. The Delta analyzer demonstrates improved overall system performance for a rapid 2-min measurement in As skin phantoms, such that it can be considered for use in populations exposed to arsenic. - Highlights: • Second generation portable XRF system reports lower Arsenic detection limit. • When excluded from calibration, <1 ppm error in predicted 5, 20 ppm As concentration. • Effective dose of both generations of systems in stand mode is within ICRP limit.

  11. Technique of sample preparation for analysis of gasoline and lubricating oils by X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Avila P, P.

    1990-03-01

    The X-ray fluorescence laboratory of the National Institute of Nuclear Research when not having a technique for the analysis of oils it has intended, with this work, to develop a preparation technique for the analysis of the metals of Pb, Cr, Ni, V and Mo in gasolines and oils, by means of the spectrometry by X-ray fluorescence analysis. The obtained results, its will be of great utility for the one mentioned laboratory. (Author)

  12. Optimization of a spectrometry for energy-dispersive x-ray fluorescence analysis by x-ray tube in combination with secondary target for multielements determination of sediment samples

    International Nuclear Information System (INIS)

    Zaidi Embong; Husin Wagiran

    1997-01-01

    The design of an energy-dispersive X-ray fluorescence spectrometer equipped with a conventional X-ray tube and secondary target is described. The spectrometer system constructed in our laboratory consists of a semiconductor detector system, irradiation chamber and X-ray tube. Primary source from X-ray tube was used to produced secondary X-ray from selenium, molybdenum and cadmium targets. The fluorescence X-ray from the sample was detected using Si(Li) detector with resolution of 0. 175 keV (Mn-K(x). The spectrometer was used for determination of multi-elements with atomic number between 20 to 44 in river sediment samples. The X-ray spectrum, from the samples were analysed using computer software which was developed based on Marquardt method. Optimal conditions and detection limits are determined experimentally by variation of excitation parameters for each combination of secondary target and tube voltage

  13. Application of X-ray fluorescence in investigation of historical monuments

    International Nuclear Information System (INIS)

    Cechak, Tomas; Trojek, Tomas; Musilek, Ladislav; Paulusova, Hana

    2008-01-01

    Full text: Nuclear techniques represent invaluable tools in non-destructive diagnostics applied to archaeological findings and objects of arts, mainly for dating and determining the composition of materials used in the production of artefacts. In this work we present the application of X-ray fluorescence analysis (XRFA). The X-ray fluorescence apparatus built and operated in the Laboratory of Quantitative Methods in Research of Ancient Monuments, constituent part of the Department of Dosimetry and Application of Ionizing Radiation, FNSPE, was used for the purpose of old relics measurements. The X-ray sources (radionuclides) generate the characteristic X-ray photons from the sample. After processing the resulting signal, we obtain information about the chemical composition of the tested sample. These results give data for qualitative and quantitative analysis of samples. XRFA is relatively simple and non-destructive method. Capability of in-situ measurement is one of big advantages of this method. The radionuclide sources of exciting radiation (e.g. 55 Fe enables the excitation of elements with Z up to 23, 238 Pu is used in interval of Z from 20 to 39 etc.) and X-ray tube with Mo anode were used. Narrow collimation of the exciting beam makes it possible to select the measured area of e.g. pigments in old manuscripts. X-ray fluorescence analysis, both in its energy form and in its wave dispersive form, is one of the most widespread methods using ionising radiation to study the elemental composition of materials. It is frequently used for studies of various cultural and historic relicts and objects of art. This work summarizes the author's experience with X-ray fluorescence analysis in investigating historical relicts namely by means of portable spectroscopic devices. Utilization of these methods is demonstrated in the investigation of fresco paintings, metal objects and old manuscripts. The results of these measurements provide the information on the composition of

  14. Analysis of Thousands of Prehistoric Mediterranean Obsidian Artifacts Using a Nondestructive Portable X-Ray Fluorescence Spectrometer

    Science.gov (United States)

    Tykot, Robert

    A portable, hand-held X-ray fluorescence spectrometer has been used for a decade to elementally analyze prehistoric obsidian artifacts in the Mediterranean. Nearly 400 geological obsidian samples and 7500 obsidian artifacts have been analyzed. The pXRF can distinguish all individual sources, as well as assign artifacts specifically to most subsources. For the island sources of Lipari, Pantelleria, Sardinia, and Melos, it is important to address the usage of obsidian from specific subsources due to human selection based on physical properties of the raw material and their production practices, which may have changed over time from the Early Neolithic to the Bronze Age. The analysis of 50 or more artifacts from 60 different archaeological sites allows for statistical comparison between sites, and their contexts, geographic areas (e.g. coastal/inland, highland/lowland) and distance from geological sources. The frequency of transport between island sources and mainland sites is suggestive of maritime capabilities also for the transport of domesticated animals, ceramics, and other materials. This presentation will specifically address potential limitations of the portable XRF, including non-destructive surface analysis of potentially heterogeneous materials, and limited trace element detection compared to other analytical methods, versus its highly beneficial ``package'' of analyzing great numbers of artifacts non-destructively and rapidly without needing to export them from museums and facilities in many countries.

  15. The use of swept-charge devices in planetary analogue X-ray fluorescence studies

    International Nuclear Information System (INIS)

    Walker, T E; Smith, D R

    2012-01-01

    The Chandrayaan-1 X-ray Spectrometer (C1XS) was launched onboard the Indian Space Research Organisation (ISRO) Chandrayaan-1 lunar mission in October 2008. The instrument consisted of 24 swept-charge device (SCD) silicon X-ray detectors providing a total collecting area of ∼ 24 cm 2 , corresponding to a 14° field of view (FWHM), with the ability to measure X-rays from 0.8–10 keV. One algorithm used to analyse the C1XS flight data was developed at Rutherford Appleton Laboratory (RAL) to convert the raw X-ray flux data into elemental ratios and abundances to make geological interpretations about the lunar surface. Laboratory X-ray fluorescence (XRF) data were used to validate the RAL algorithm, with previous studies investigating how the measured XRF flux varies with target surface characteristics including grain size and roughness. Evidence for a grain-size effect was observed in the data, the XRF line intensity generally decreasing with increasing sample grain size, dependent on the relative abundance of elemental components. This paper presents a subsequent study using more homogeneous samples made from mixtures of MgO, Al 2 O 3 and SiO 2 powders, all of grain size < 44 μm, across a broader range of mixture ratios and at a higher level of X-ray flux data in order to further validate the RAL algorithm. For the majority of the C1XS flight data analysed so far with the RAL algorithm, the corresponding lunar ground tracks have been generally basaltic, laboratory verification of the algorithm having been primarily conducted using basaltic lunar regolith simulant (JSC-1A) XRF data. This paper also presents results from tests on a terrestrial anorthosite sample, more relevant to the anorthositic lunar highlands, from where the remaining C1XS lunar dataset derives. The operation of the SCD, the XRF test facility, sample preparation and collected XRF spectra are discussed in this paper.

  16. Uncertainty calculations for the measurement of in vivo bone lead by x-ray fluorescence

    International Nuclear Information System (INIS)

    O'Meara, J M; Fleming, D E B

    2009-01-01

    In order to quantify the bone lead concentration from an in vivo x-ray fluorescence measurement, typically two estimates of the lead concentration are determined by comparing the normalized x-ray peak amplitudes from the Kα 1 and Kβ 1 features to those of the calibration phantoms. In each case, the normalization consists of taking the ratio of the x-ray peak amplitude to the amplitude of the coherently scattered photon peak in the spectrum. These two Pb concentration estimates are then used to determine the weighted mean lead concentration of that sample. In calculating the uncertainties of these measurements, it is important to include any covariance terms where appropriate. When determining the uncertainty of the lead concentrations from each x-ray peak, the standard approach does not include covariance between the x-ray peaks and the coherently scattered feature. These spectral features originate from two distinct physical processes, and therefore no covariance between these features can exist. Through experimental and simulated data, we confirm that there is no observed covariance between the detected Pb x-ray peaks and the coherently scattered photon signal, as expected. This is in direct contrast to recent work published by Brito (2006 Phys. Med. Biol. 51 6125-39). There is, however, covariance introduced in the calculation of the weighted mean lead concentration due to the common coherent normalization. This must be accounted for in calculating the uncertainty of the weighted mean lead concentration, as is currently the case. We propose here an alternative approach to calculating the weighted mean lead concentration in such a way as to eliminate the covariance introduced by the common coherent normalization. It should be emphasized that this alternative approach will only apply in situations in which the calibration line intercept is not included in the calculation of the Pb concentration from the spectral data: when the source of the intercept is well

  17. Application of radionuclide sources for excitation in energy-dispersive X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Hoffmann, P.

    1986-01-01

    X-ray fluorescence (XRF) analysis is in broad application in many fields of science where elemental determinations are necessary. Solid and liquid samples are analyzed by this method. Solids are introduced in thin or thick samples as melted glass, pellets, powders or as original specimen. The excitation of X-ray spectra can be performed by specific and polychromic radiation of X-ray tubes, by protons, deuterons, α-particles, heavy ions and synchrotron radiation from accelerators and by α-particles, X- and γ-rays and by bremsstrahlung generated by β - -particles from radionuclide sources. The radionuclides are devided into groups with respect to their decay mode and the energy of the emitted radiation. The broad application of radionuclides in XRF excitation is shown in examples as semi-quantitative analysis of glasses, as quantitative analysis of coarse ceramics and as quantitative determination of heavy elements (mainly actinides) in solutions. The advantages and disadvantages of radionuclide excitation in XRF analysis are discussed. (orig.) [de

  18. TX 2000: total reflection and 45o energy dispersive x-ray fluorescence spectrometer

    International Nuclear Information System (INIS)

    Pasti, F.; Torboli, A.; Valdes, M.

    2000-01-01

    This equipment, developed by Ital Structures, combines two kinds of energy dispersive X-ray fluorescence techniques, the first using total reflection geometry and the second conventional 45 o geometry. The equipment is completely controlled by a PC and to reach the condition of total reflection is very easy because it is enough to load the file with the right position for the corresponding energy. In this apparatus we used an x-ray tube with an alloy anode of Mo/W with a long fine focus at 2200 W. To monochromatize the x-ray beam while choosing, for example, the Mo K alpha or W L alpha or a piece of white spectrum of 33 keV, we use a highly reflective multilayer made of Si/W with 2d = 45.5 A o . The detector used in the equipment is a lithium drifted silicon detector (Si(Li)) with an excellent energy resolution of 135 eV at 5.9 keV and 1000 cps. We developed two programs written in Windows 95, 98 and NT for a 32 bit microprocessor. The first one is called TYACQ32 and has the following functions: first, complete control of the hardware, second automatic alignment of the TX 2000 spectrometer and third acquisition of spectra. The second program is EDXRF32. This is a program to accomplish spectrum and quantitative analysis for TXRF and EDXRF 45 o degrees analysis. (author)

  19. Methodology for determination of inorganic elements in vitamin supplements by X-ray fluorescence

    International Nuclear Information System (INIS)

    Endo, Lucas Y.; Portezan Filho, Otavio; Perreira, Paulo S.; Lopes, Fabio; Appoloni, Carlos R.; Melquiades, Fabio L.

    2009-01-01

    A methodology is present for qualification of inorganic elements in alimentary supplements, using a portable system of low power X-ray, in the analysis of two commercial products, both found as pills. The PXRF (Portable X-ray Fluorescence) is a X-ray mini tube, Ag filter at tube exit, a Si-Pin semiconductor and a Ag collimator at the detector entry, specific circuitry for this type of system and a microcomputer for the data acquisition and analysis. Samples and standards were diluted in boric acid, and the samples were prepared in a 1:10 rate, and the standards vary of 1:2, 1;5, 1:10. 1:15 and 1:100. Results were obtained for Ca, Ti, Mn, Fe, Cu e Zn, with values varying from 0.14 to 8.45 for the sample A, and from 0.17 to 17.94 for the sample B. The PXRF portable system demonstrated to be a multi elementary analysis tool with high analytical speed, and can be applied in situ no only for quality control but also for monitoring of inspection organism

  20. X-ray fluorescence and computed radiography analysis of a famous brazilian painting from XIX century

    International Nuclear Information System (INIS)

    Calza, Cristiane; Oliveira, Davi F.; Rocha, Henrique S.; Lopes, Ricardo Tadeu

    2009-01-01

    This work used Energy Dispersive X-Ray Fluorescence (EDXRF) and Computed Radiography (CR) to evaluate the general conditions of the painting 'Gioventu' (Eliseu Visconti, 1898), identifying possible problems, areas that revealed signs of previous retouchings and the pigments used by the artist. EDXRF measurements were carried out with a portable system consisting of an X-ray tube Oxford TF3005 with W anode, operating at 25 kV and 100 μA, and a Si-PIN XR-100CR detector from Amptek. Several spectra were obtained in each color, with an acquisition time of 300 s and a beam collimation of 2 mm. The spectra were analyzed using the software QXAS-AXIL (IAEA). The results revealed that the drawings were made over a preparatory layer of lead white. Some pigments identified were: yellow, red and brown ochre; umbra; vermilion; cobalt blue; etc. The experimental setup used in the CR analysis consisted of an Oxford X-ray source, operating at 50 kV and 200 μA, placed at 85 cm from the painting, a GE CR 50P portable computed radiography scanner and a Fuji imaging plate detector. The exposure time was 600 s. The radiographic images revealed that the painting was in a good state of conservation and also a complete composition hidden underneath the visible paint layer.(author)

  1. Total reflection X-ray Fluorescence determination of interfering elements rubidium and uranium by profile fitting

    Science.gov (United States)

    Dhara, Sangita; Khooha, Ajay; Singh, Ajit Kumar; Tiwari, M. K.; Misra, N. L.

    2018-06-01

    Systematic studies to assess the analytical parameters obtained in the total reflection X-ray fluorescence (TXRF) determinations of interfering elements Rb and U using profile fitting are reported in the present manuscript. The X-ray lines Rb Kα and U Lα having serious spectral interference (ΔE = 218 eV), have been used as analytical lines. The intensities of these X-ray lines have been assessed using profile fitting. In order to compare the analytical results of Rb determinations in presence of U, with and without U excitation, synchrotron radiation was tuned to energy just above and below the U Labs edge. This approach shall excite both Rb Kα and U Lα simultaneously and Rb Kα selectively. Finally, the samples were also analyzed with a laboratory based TXRF spectrometer. The analytical results obtained in all these conditions were comparable. The authenticity of the results was assessed by analyzing U with respect to Rb in Rb2U(SO4)3, a standard reference material for U. The average precision obtained for TXRF determinations was below 3% (RSD, n = 3, 1σ) and the percent deviation of TXRF values from the expected values calculated on the basis of sample preparation was within 3%.

  2. XRF Newsletter, No. 25, September 2013 [X Ray Fluorescence in the IAEA and its Member States

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2014-02-15

    One of the main missions of any analytical laboratory is the provision of reliable results. Many X-ray fluorescence laboratories in the International Atomic Energy Agency (IAEA) Member States (MSs) carry out research aimed at improving the performance and extending the applicability of various X-ray based techniques. The IAEA assists its MSs Laboratories to maintain their readiness by producing reference materials, by developing standardized analytical methods and by conducting interlaboratory comparisons and proficiency tests as tools for quality control. To achieve this aim and ensure a reliable worldwide, rapid and consistent response, the IAEA Nuclear Science and Instrumentation Laboratory (NSIL) organizes proficiency tests annually for X-ray spectrometry laboratories. The main objective of these tests is to enhance the capability of interested MSs in effective utilization of nuclear spectrometry and analytical services in industry, human health, agriculture, and in monitoring and evaluation of environmental pollution. Indeed, proficiency tests are designed to identify analytical problems, to support IAEA MSs laboratories to improve the quality of their analytical results, to maintain their accreditation and to provide a regular forum for discussion and technology transfer in this area. The type of samples and the concentration levels of the analytes are designed and chosen in a way to better enable identification of potential analytical problems. Since 2001, the NSIL has organized nine proficiency test exercises (PTXRFIAEA)

  3. Radioisotope induced energy dispersive X-ray fluorescence - a diagnostic tool in clinical science

    International Nuclear Information System (INIS)

    Joseph, Daisy

    2010-01-01

    Full text: Energy dispersive X-ray fluorescence (EDXRF) - an ideal technique for detecting trace elements in drugs have been used for analyzing drugs marked as Zn supplements (Jasad Bhasm) used for growth in children and Ayurvedic medicines containing toxic elements such as Arsenic (As) and Mercury (Hg). Folklore medicines obtained as plants extracts from Manipur plants were also analyzed for their composition. Zn supplements (Jasad Bhasm) manufactured by various manufacturers were analyzed for their trace elements besides Zn and were compared with laboratory preparations. Similarly the Ayurvedic medicines from different companies were analyzed for their metal composition. All samples in powder form were pelletized and analyzed using an X-ray spectrometer consisting of a Cd 109 radioisotope source, Si (Li) detector of resolution 170 eV at 5.9 KeV Mn X-ray, preamplifier, amplifier and a PC based multichannel analyzer. Varying amounts of trace elements were detected in Jasad Bhasm and interesting results (As and Hg) were seen in the Ayurvedic medicines in addition to other trace elements such as K,Ca,Fe,Cu and Zn. In Manipur plant extracts Sr was predominantly seen in most samples. Their levels of toxicity and significance to human health and diseases will be discussed in the remaining sections of the paper

  4. Development of Ta-based Superconducting Tunnel Junction X-ray Detectors for Fluorescence XAS

    International Nuclear Information System (INIS)

    Friedrich, S.; Drury, O.; Hall, J.; Cantor, R.

    2009-01-01

    We are developing superconducting tunnel junction (STJ) soft X-ray detectors for chemical analysis of dilute samples by fluorescence-detected X-ray absorption spectroscopy (XAS). Our 36-pixel Nb-based STJ spectrometer covers a solid angle (Omega)/4π ∼ 10 -3 , offers an energy resolution of ∼10-20 eV FWHM for energies up to ∼1 keV, and can be operated at total count rates of ∼10 6 counts/s. For increased quantum efficiency and cleaner response function, we have now started the development of Ta-based STJ detector arrays. Initial devices modeled after our Nb-based STJs have an energy resolution below 10 eV FWHM for X-ray energies below 1 keV, and pulse rise time discrimination can be used to improve their response function for energies up to several keV. We discuss the performance of the Ta-STJs and outline steps towards the next-generation of large STJ detector arrays with higher sensitivity.

  5. X-ray fluorescence and computed radiography analysis of a famous brazilian painting from XIX century

    Energy Technology Data Exchange (ETDEWEB)

    Calza, Cristiane; Oliveira, Davi F.; Rocha, Henrique S.; Lopes, Ricardo Tadeu [Coordenacao dos Programas de Pos-graduacao de Engenharia (COPPE/UFRJ), Rio de Janeiro, RJ (Brazil). Lab. de Instrumentacao Nuclear], e-mail: ccalza@lin.ufrj.br; Pedreira, Andrea [Museu Nacional de Belas Artes, Rio de Janeiro, RJ (Brazil)

    2009-07-01

    This work used Energy Dispersive X-Ray Fluorescence (EDXRF) and Computed Radiography (CR) to evaluate the general conditions of the painting 'Gioventu' (Eliseu Visconti, 1898), identifying possible problems, areas that revealed signs of previous retouchings and the pigments used by the artist. EDXRF measurements were carried out with a portable system consisting of an X-ray tube Oxford TF3005 with W anode, operating at 25 kV and 100 {mu}A, and a Si-PIN XR-100CR detector from Amptek. Several spectra were obtained in each color, with an acquisition time of 300 s and a beam collimation of 2 mm. The spectra were analyzed using the software QXAS-AXIL (IAEA). The results revealed that the drawings were made over a preparatory layer of lead white. Some pigments identified were: yellow, red and brown ochre; umbra; vermilion; cobalt blue; etc. The experimental setup used in the CR analysis consisted of an Oxford X-ray source, operating at 50 kV and 200 {mu}A, placed at 85 cm from the painting, a GE CR 50P portable computed radiography scanner and a Fuji imaging plate detector. The exposure time was 600 s. The radiographic images revealed that the painting was in a good state of conservation and also a complete composition hidden underneath the visible paint layer.(author)

  6. Multilayers quantitative X-ray fluorescence analysis applied to easel paintings.

    Science.gov (United States)

    de Viguerie, Laurence; Sole, V Armando; Walter, Philippe

    2009-12-01

    X-ray fluorescence spectrometry (XRF) allows a rapid and simple determination of the elemental composition of a material. As a non-destructive tool, it has been extensively used for analysis in art and archaeology since the early 1970s. Whereas it is commonly used for qualitative analysis, recent efforts have been made to develop quantitative treatment even with portable systems. However, the interpretation of the results obtained with this technique can turn out to be problematic in the case of layered structures such as easel paintings. The use of differential X-ray attenuation enables modelling of the various layers: indeed, the absorption of X-rays through different layers will result in modification of intensity ratio between the different characteristic lines. This work focuses on the possibility to use XRF with the fundamental parameters method to reconstruct the composition and thickness of the layers. This method was tested on several multilayers standards and gives a maximum error of 15% for thicknesses and errors of 10% for concentrations. On a painting test sample that was rather inhomogeneous, the XRF analysis provides an average value. This method was applied in situ to estimate the thickness of the layers a painting from Marco d'Oggiono, pupil of Leonardo da Vinci.

  7. Quantification of phosphorus in single cells using synchrotron X-ray fluorescence

    Energy Technology Data Exchange (ETDEWEB)

    Núñez-Milland, Daliángelis R. [Department of Chemistry and Biochemistry, University of South Carolina, Columbia, SC 29208 (United States); Baines, Stephen B. [Department of Ecology and Evolution, Stony Brook University, Stony Brook, NY 11755 (United States); Vogt, Stefan [Experimental Facilities Division, Advanced Photon Source, Argonne National Laboratory, Argonne, IL (United States); Twining, Benjamin S., E-mail: btwining@bigelow.org [Department of Chemistry and Biochemistry, University of South Carolina, Columbia, SC 29208 (United States)

    2010-07-01

    Phosphorus abundance was quantified in individual phytoplankton cells by synchrotron X-ray fluorescence and compared with bulk spectrophotometric measurements to confirm accuracy of quantification. Figures of merit for P quantification on three different types of transmission electron microscopy grids are compared to assess possible interferences. Phosphorus is required for numerous cellular compounds and as a result can serve as a useful proxy for total cell biomass in studies of cell elemental composition. Single-cell analysis by synchrotron X-ray fluorescence (SXRF) enables quantitative and qualitative analyses of cell elemental composition with high elemental sensitivity. Element standards are required to convert measured X-ray fluorescence intensities into element concentrations, but few appropriate standards are available, particularly for the biologically important element P. Empirical P conversion factors derived from other elements contained in certified thin-film standards were used to quantify P in the model diatom Thalassiosira pseudonana, and the measured cell quotas were compared with those measured in bulk by spectrophotometry. The mean cellular P quotas quantified with SXRF for cells on Au, Ni and nylon grids using this approach were not significantly different from each other or from those measured spectrophotometrically. Inter-cell variability typical of cell populations was observed. Additionally, the grid substrates were compared for their suitability to P quantification based on the potential for spectral interferences with P. Nylon grids were found to have the lowest background concentrations and limits of detection for P, while background concentrations in Ni and Au grids were 1.8- and 6.3-fold higher. The advantages and disadvantages of each grid type for elemental analysis of individual phytoplankton cells are discussed.

  8. A strategy for vitrification product assurance and control with X-ray fluorescence spectrometry (PACX)

    International Nuclear Information System (INIS)

    Resce, J.L.

    1995-01-01

    A product control strategy is proposed for the vitrification of low-level and/or mixed waste. This strategy is called Product Assurance and Control with X-ray fluorescence spectrometry (PACX). The strategy utilizes sequential wavelength dispersive x-ray fluorescence spectrometry and standardless fundamental parameters calculations to analyze both the melter feed and the glassy products. The melter feed is sampled prior to addition to the melt tank and then melted and cast into samples which should closely resemble the product from the vitrification process itself. The resulting sample disks are then analyzed by x-ray fluorescence spectrometry. All elements with atomic numbers down to sodium can be determined directly and then either boron or lithium can be determined by difference from the mass balance. The XRF intensities are converted into oxide compositions with the use of a novel standardless fundamental parameters program. Previous work has shown that there is an excellent correlation between the XRF results and the results from conventional wet chemical analyses, but the XRF results can be obtained within two to three hours of sampling. If compositional control limits for durability are available, the product acceptability can be determined prior to the batch being introduced into the melter. The durability could also be estimated from a model, if available, which predicts product durability from composition. If the predicted durability is estimated to be too low, the model can then be used to determine additives which will raise the durability of the final product to within acceptable limits. The additives can then be incorporated into the batch prior to addition into the melter. A similar XRF analysis can be carried out on the glass product from the melter which can then be used to predict and thus assure product acceptance

  9. Sample preparation of waste water to determine metallic contaminants by X-ray fluorescence spectroscopy

    International Nuclear Information System (INIS)

    Gonzalez Olivos, Javier.

    1987-01-01

    Trace X-ray fluorescence spectroscopy analysis in liquid samples is preceded by sample preparation, which usually consists in the precipitation of the metallic ions and concentration over a thin cellulose filter. The samples preparation of waste water by this method is not efficient, due to the great amount of organic and insoluble matter that they contain. The purpose of this work was to determine the optimal value of pH in order to adsorbe all the insoluble matter contained in a waste water sample in the activated charcoal, so that the metallic ions could be precipitated and concentrated on a thin filter and determinated by X-ray fluorescence spectroscopy. A survey about the adsorption of some ions in activated charcoal in function of the pH was made for the following: Cr 3+ , Fe 3+ , Ni 2+ , Cu 2+ , Zn 2+ , Se 2+ , Hg 2+ , and Pb 2+ . It was observed that at pH 0, the ions are not adsorbed, but Cu 2+ and Zn 2+ are adsorbed in small amount; at pH 14, the ions are adsorbed, excluding Se, which is not adsorbed at any value of pH. If a waste water sample is treated at pH 0 with activated charcoal to adsorbe the organic and insoluble matter, most of the metallic ions are not adsorbed by the activated charcoal and could be precipitated with APDC (ammonium 1-pirrolidine dithio carbamate salt) and concentrated on a thin filter. The analysis of the metallic ions contained on the filter and those adsorbed in the activated charcoal by X-ray fluorescence spectroscopy, gave the total amount of the ions in the sample. (author)

  10. Total reflection X-ray fluorescence and archaeometry: Application in the Argentinean cultural heritage

    Energy Technology Data Exchange (ETDEWEB)

    Vazquez, Cristina [Comision Nacional de Energia Atomica, Av. Gral Paz 1499 (B1650KNA) San Martin, Buenos Aires (Argentina); Laboratorio de Quimica de Sistemas Heterogeneos, Facultad de Ingenieria, Universidad de Buenos Aires, P. Colon 850 (C1063ACU), Buenos Aires (Argentina)], E-mail: Cristina.Vazquez@cnea.gov.ar; Albornoz, Ana [Agencia Rio Negro Cultura, Museo de la Patagonia F.P.Moreno, Centro Civico s/n Bariloche, Rio Negro (Argentina); Hajduk, Adam [CONICET, Museo de la Patagonia F.P.Moreno, Centro Civico s/n Bariloche, Rio Negro (Argentina); Elkin, Dolores [CONICET Instituto Nacional de Antropologia y Pensamiento Latinoamericano, 3 de febrero 1378 (C1426AEL) Buenos Aires (Argentina); Custo, Graciela; Obrustky, Alba [Comision Nacional de Energia Atomica, Av. Gral Paz 1499 (B1650KNA) San Martin, Buenos Aires (Argentina)

    2008-12-15

    Archaeometry is an interdisciplinary research area involved in the development and use of scientific methods in order to answer questions concerned with the human history. In this way the knowledge of archaeological objects through advanced chemical and physical analyses permits a better preservation and conservation of the cultural heritage and also reveals materials and technologies used in the past. In this sense, analytical techniques play an important role in order to provide chemical information about cultural objects. Considering the non destructive characteristic of this study, analytical techniques must be adequate in order to prevent any alteration or damage and in addition to allow the conservation of their integrity. Taking into account the irreplaceable character of the archaeological and artistic materials considered in this study, analytical techniques must be adequate in order to prevent any alteration or damage and in addition to allow the conservation of their integrity. Total Reflection X-ray Fluorescence Spectrometry as a geometric variant of conventional X-ray fluorescence is a proved microanalytical technique considering the small amount of sample required for the analysis. A few micrograms are enough in order to reveal valuable information about elemental composition and in this context it is highly recommended for artwork studies. In this paper a case study is presented in which Total Reflection X-Ray Fluorescence Spectrometry has been successfully employed in the archaeometry field. Examples from Argentinean cultural heritage sites related with the determination of pigments in paintings on canvas and in rock sites as well as in underwater archaeology research are shown.

  11. Topics: in vivo measurement of thyroidal iodine content by x-ray fluorescent technique

    International Nuclear Information System (INIS)

    Imamura, Keiko

    1979-01-01

    Thyroidal iodine content gives useful informations in the fields of physiology, clinical medicine, health physics etc. Iodine content has been determined mainly for resected thyroids. Recently, x-ray fluorescent analysis has been extended as the in vivo technique first in the clinical medicine. Exciting sources used for the analysis of the thyroid are Am-241 or x-ray tube. Am-241 has a half-life of 438 years and emits #betta#-ray of 60 keV. Thyroid can be imaged by fluorescent scan utilizing strong (10 - 15 Ci) Am-241 source. Examination time is about 15 min and the radiation dose to the gland is about 15 - 60 mrad. Iodine content is determined by static fluorescent technique equipped with weaker source of less than 1 Ci. Thyroidal iodine content in normal subjects were analysed by this technique and the results were in good accordance with those obtained by in vitro analysis. Difference in the thyroidal iodine content between the Japanese population and other countries is not clear. Application to the pathological cases has provided many findings about the iodine content and its distribution which could not be obtained by in vitro analysis. This in vivo technique can be safely performed for infants and for pregnancies, and the relatively compact size of this apparatus could be widely used in the study of health physics and environmental problems. (author)

  12. Determination of coal ash content by the combined x-ray fluorescence and scattering spectrum

    Science.gov (United States)

    Mikhailov, I. F.; Baturin, A. A.; Mikhailov, A. I.; Borisova, S. S.; Fomina, L. P.

    2018-02-01

    An alternative method is proposed for the determination of the inorganic constituent mass fraction (ash) in solid fuel by the ratio of Compton and Rayleigh X-ray scattering peaks IC/IR subject to the iron fluorescence intensity. An original X-ray optical scheme with a Ti/Mo (or Sc/Cu) double-layer secondary radiator allows registration of the combined fluorescence-and-scattering spectrum at the specified scattering angle. An algorithm for linear calibration of the Compton-to-Rayleigh IC/IR ratio is proposed which uses standard samples with two certified characteristics: mass fractions of ash (Ad) and iron oxide (WFe2O3). Ash mass fractions have been determined for coals of different deposits in the wide range of Ad from 9.4% to 52.7% mass and WFe2O3 from 0.3% to 4.95% mass. Due to the high penetrability of the probing radiation with energy E > 17 keV, the sample preparation procedure is rather simplified in comparison with the traditional method of Ad determination by the sum of fluorescence intensities of all constituent elements.

  13. High temperature monitoring of silicon carbide ceramics by confocal energy dispersive X-ray fluorescence spectrometry

    Energy Technology Data Exchange (ETDEWEB)

    Li, Fangzuo; Liu, Zhiguo; Sun, Tianxi, E-mail: stx@bnu.edu.cn

    2016-04-15

    Highlights: • X-ray scattering was used for monitoring oxidation situation of SiC ceramics. • A calibration curve was obtained. • The confocal X-ray scattering technology was based on polycapillary X-ray optics. • The variations of contents of components of SiC ceramics were obtained. - Abstract: In the present work, we presented an alternative method for monitoring of the oxidation situation of silicon carbide (SiC) ceramics at various high temperatures in air by measuring the Compton-to-Rayleigh intensity ratios (I{sub Co}/I{sub Ra}) and effective atomic numbers (Z{sub eff}) of SiC ceramics with the confocal energy dispersive X-ray fluorescence (EDXRF) spectrometer. A calibration curve of the relationship between I{sub Co}/I{sub Ra} and Z{sub eff} was established by using a set of 8 SiC calibration samples. The sensitivity of this approach is so high that it can be easily distinguished samples of Z{sub eff} differing from each other by only 0.01. The linear relationship between the variation of Z{sub eff} and the variations of contents of C, Si and O of SiC ceramics were found, and the corresponding calculation model of the relationship between the ΔZ and the ΔC{sub C}, ΔC{sub Si}, and ΔC{sub O} were established. The variation of contents of components of the tested SiC ceramics after oxidation at high temperature was quantitatively calculated based on the model. It was shown that the results of contents of carbon, silicon and oxygen obtained by this method were in good agreement with the results obtained by XPS, giving values of relative deviation less than 1%. It was concluded that the practicality of this proposed method for monitoring of the oxidation situation of SiC ceramics at high temperatures was acceptable.

  14. Analysis of agricultural soils by using energy dispersive X-ray fluorescence

    International Nuclear Information System (INIS)

    Anjos, Marcelino Jose dos

    2000-03-01

    In this work, we describe an Energy Dispersive x-ray Fluorescence System with a x-ray tube excitation for trace analysis of environmental samples (soil). The system was used to analyze the contamination of metals in treated soils with doses of 10, 20 and 30 ton/ha of compound organic of urban garbage of the type Fertilurb and 10 ton/ha of aviary bed (manure of birds). Samples of roots and foliages of plant radishes cultivated in these soils were also analyzed. The soil samples were collected in five different depths of 0,5, 5-10, 10-20, 20-40 and 40-60 cm. The experimental set-up is composed by an OXFORD X-ray (30 kV, 50 μA and W anode), an ORTEC Si-Li detector, with an energy resolution of about 180 eV at 5.9 keV and an ORTEC multichannel-analyser. The X-ray spectrum tube is quasi-monochromatic by using of Ti filter. Samples were prepared in pellet form with superficial density in the range of 100 mg/cm 2 . The fundamental parameter method was used in order to verify the elemental concentration. It was possible to determine the concentrations of thirteen elements: K, Ca, Ti, Mn, Fe, Cu, Zn, Br, Rb, Sr, Y, Zr and Pb in the treated soils with compounds organic. The results indicate that the values found for K, Ca, Rb, Sr, Zr and Pb are significantly above the upper confidence limits for the control soil (α = 0.05). There is a real different between these elements compared to their relationship in the control soils, (α=0,05). There is a real difference between these elements compared to their relationship in the control soils, confirming the influence of the organic compounds in the soil. (author)

  15. Wavelength Dispersive X-ray Fluorescence Spectrometry for the Analysis of Organic Polymer Film

    International Nuclear Information System (INIS)

    Choi, Yong Suk; Park, Yong Joon; Kim, Jong Yun

    2008-01-01

    Recently, many studies have been focused on the thin films because there are numerous industrial processes relevant to thin films such as fuel cells, sensors, lubricants, coatings, and so on. Physical and chemical properties of solid surface have been modified by ultra-thin coatings such as Langmuir-Blodgett (LB) method with a variety of types of organic functional materials for the specific purposes in many applications. In addition, the layer-by-layer technique using polyelectrolyte films are now of interest as biosensors, electrochromic and electroluminescent devices, etc. In general, several methods such as X-ray or neutron reflectivity, and quartz crystal microbalance (QCM) have been utilized for the thin film analysis. These optical techniques can measure the film thicknesses up to hundreds of nanometers while X-ray photoelectron spectroscopy is widely used to study a few nanometers thick films. Other methods such as X-ray Photoelectron spectroscopy (XPS), scanning electron microscopy (SEM), transmission electron microscopy (TEM), and atom force microscopy (AFM) have also been used in the film analysis in spite of some disadvantages for each method. X-ray fluorescence (XRF) has long been used as a rapid and simple analytical tool for the analysis of elemental composition of materials. XRF technique is suitable for on-line or in-line real-time monitoring because it is a non-destructive and rapid analysis with good precision and good accuracy at low cost. The aim of this work is to develop a new analytical technique for the quantitative analysis of polymer film on metal substrate. In the present study, Compton peak profile was investigated under different experimental conditions by using wavelength-dispersive XRF (WD-XRF). Compared to energy-dispersive XRF (ED-XRF), WD-XRF is more adequate in an accurate quantitative analysis of thin organic film

  16. Investigations of some regional river systems by INAA and X-ray fluorescence

    International Nuclear Information System (INIS)

    Drazhkovich, R.J.; Kukoch, A.

    1985-01-01

    Distribution of Zn, Hg, Sb, Cr, Fe, Sc and Co has been investigated in materials dissolved and suspended in the rivers Ibar, Zapadna Morava and Kamenica by means of INAA and X-ray fluorescence. Irradiation was made in VKG-channels of RA-nuclear reactor Vincha. Distribution coefficients were calculated, as well as contamination factors for investigated river regional systems in comparison to the uncontaminated water system. Data obtained indicate the possibility of utilization of these two analytical methods for investigation and control of biogeochemical and contamination processes in small regional water systems, especially important for modern studies in life sciences

  17. X-ray fluorescence diffractionless analyzer for determining light element content in iron ore mixtures

    International Nuclear Information System (INIS)

    Yuksa, L.K.; Kochmola, N.M.; Bondarenko, V.P.; Bogdanov, V.K.

    1986-01-01

    Diffractionless X-ray fluorescence analyzer for detecting calcium oxide and silicon dioxide contents in dry iron ore materials has been developed. The analyzer includes a charging unit, sample-conveying device, spectrometric units for detecting calcium and silicon, computing racks and sample-removing device. Results of calcium oxide and silicon dioxide analyses in iron ore mixtures are presented. Errors are evaluated. It is shown that the analyzer provides high accuracy of one-time determinations, as well as reading constancy for a long time

  18. Heavy metals analysis in blood by the X-ray fluorescence method

    International Nuclear Information System (INIS)

    Perez Novara, Ana Ma.

    1988-05-01

    The analytical procedure for determination of heavy metals in blood is described. Blood was taken from active smelter workers, samples were stored in special flasks at 2 grades C. After freeze drying, dried samples was analyzed using X-ray fluorescence, sources excited: Pu-238 and Cd-109 was used. Br in Pb interferences were corrected. The agreement of the results with values in similar workers is satisfactory Median Pb level was 42.6 micro grams/100 ml (total blood), Cu was less 2.5 micro grams/g (dry blood) and Zn was 11 micro grams/g (dry blood). (author)

  19. The capabilities of total reflection X-ray fluorescence in the polymeric analytical field

    International Nuclear Information System (INIS)

    Vazquez, Cristina

    2004-01-01

    This paper presents the capabilities of total reflection X-ray fluorescence (TXRF) as analytical technique specially focused in high-viscosity polymer dispersions. Appropriate sample preparation procedures are described taking into account the time stability of these dispersions. Special remarks considering different ways for drying samples in order to obtain the most uniform deposited film are investigated focusing on the behavior of aqueous and viscous systems. Due to the difficulty found for obtaining a homogeneous mixture between sample and internal standard in such systems, the use of Compton incoherent scattering is discussed as an alternative procedure for trace quantification in high viscous systems without any mixing processes

  20. The capabilities of total reflection X-ray fluorescence in the polymeric analytical field*1

    Science.gov (United States)

    Vázquez, Cristina

    2004-08-01

    This paper presents the capabilities of total reflection X-ray fluorescence (TXRF) as analytical technique specially focused in high-viscosity polymer dispersions. Appropriate sample preparation procedures are described taking into account the time stability of these dispersions. Special remarks considering different ways for drying samples in order to obtain the most uniform deposited film are investigated focusing on the behavior of aqueous and viscous systems. Due to the difficulty found for obtaining a homogeneous mixture between sample and internal standard in such systems, the use of Compton incoherent scattering is discussed as an alternative procedure for trace quantification in high viscous systems without any mixing processes.

  1. Use of X-ray fluorescence for metal determination in polymers

    International Nuclear Information System (INIS)

    Guidorizzi, Lorenza

    1996-01-01

    X-Ray fluorescence spectrometry was used to determine metals and non-metals in polyester polymers. The greatest advantage of this technique over others like Atomic Absorption or Plasma Emission is that no sample previous treatment (like calcination or acid digestion) is required. Other advantage of this method is its fastness allowing a complete analysis in just few minutes. On the other hand, this method requires metals higher than 15 ppm. Below those values there is a loss of the analysis' precision. Another advantage of this technique is the possibility of making qualitative metal analysis, scanning unknown samples and identifying the found peaks automatically. (author)

  2. Fast neutron activation analysis and radioisotope X-ray fluorescence study on KALEWA and NAMMA coal

    International Nuclear Information System (INIS)

    Naing-Win

    1981-07-01

    Kalewa coal was studied with Fast Neutron Activation Analysis (FNAA) technique, employing KAMAN A-710 neutron generator and HP(Ge) detector coupled to ''Canberra'' series 30 MCA. Sequential irradiation and dual aluminium foil monitoring method was employed. Simultaneous multielement analysis was carried out. Namma Coal was studied with radioisotope X-ray Fluorescence (XRF) technique, employing Co-57 exciter source and HP(Ge) detector coupled to ''Canberra'' series 40 MCA. In both FNAA and XRF study, the results obtained were compared to that obtained with Atomic Absorption Spectrophotometry (AAS) technique. Finally, the results were reviewed together with those obtained from similar work on coal with FNAA and XRF techniques. (author)

  3. Elemental Characterization of minerals in Chenopodium quinoa grains by the X-ray fluorescence technique

    International Nuclear Information System (INIS)

    Olivera de Lescano, Paula; Nieto Aco, Daniel

    2014-01-01

    In order to characterize by the technique of X-ray fluorescence energy dispersive quinoa grain for human consumption, 8 samples of different trademarks of quinoa, which are distributed in the local market were analyzed; together one reference materials certified by the International Atomic Energy Agency (IAEA) was analyzed. The results show the presence of elements such as K, Ca, Mn, Fe, Cu, Zn, Rb and Sr, the same as compared to data reported in various studies in neighboring countries like Ecuador, Chile and Bolivia. (authors).

  4. X-ray fluorescence control of chemical composition of cast iron

    International Nuclear Information System (INIS)

    Prekina, I.M.; Rozova, O.F.; Loran, A.V.; Teplitskaya, G.A.; Smagunova, A.N.

    1995-01-01

    A method of x-ray fluorescence analysis developed for analytical set (KRF-18 diffractometer/DVK-3 computer) is used to control cast iron composition. A quantitative evaluation of errors attributed to the violation of conditions of cast iron sampling from the flow and to the quality of preparing samples for XFA is obtained. It is shown that the main component of the integral experimental error is attributed to nonuniformity of chemical composition of cast iron. Metrological studies show that reproductibility, convergence, accuracy, and sensitivity of the method match the requirements characteristic of the control process. 4 refs.; 2 tabs

  5. Data processing of X-ray fluorescence analysis using an electronic computer

    International Nuclear Information System (INIS)

    Yakubovich, A.L.; Przhiyalovskij, S.M.; Tsameryan, G.N.; Golubnichij, G.V.; Nikitin, S.A.

    1979-01-01

    Considered are problems of data processing of multi-element (for 17 elements) X-ray fluorescence analysis of tungsten and molybdenum ores. The analysis was carried out using silicon-lithium spectrometer with the energy resolution of about 300 eV and a 1024-channel analyzer. A characteristic radiation of elements was excited with two 109 Cd radioisotope sources, their general activity being 10 mCi. The period of measurements was 400 s. The data obtained were processed with a computer using the ''Proba-1'' and ''Proba-2'' programs. Data processing algorithms and computer calculation results are presented

  6. Analysis of signal to background ratio in synchrotron radiation X-ray fluorescence

    International Nuclear Information System (INIS)

    Sakurai, Kenji; Gohshi, Yohichi; Iida, Atsuo.

    1988-01-01

    The signal to background (S/B) ratio in energy dispersive X-ray fluorescence using synchrotron radiation (SR) was quantitatively analyzed. The S/B ratio, which has been significantly improved by taking advantage of the polarized nature of SR, was found to be strongly dependent on geometrical factors of the measurement system. From the analysis on the origin of the scattered background, the dependence of the S/B ratio on the geometry was quantitatively explained, mainly by the polarization properties of SR. Experimental conditions could be optimized by adjusting the degree of polarization of the incident beam and the detector solid angle. (author)

  7. Determination of heavy metals in Damascus drinking water using total reflection x-ray fluorescence

    International Nuclear Information System (INIS)

    Bakraji, E. H.; Karajo, J.

    2000-01-01

    Total reflection x-ray fluorescence spectrometry and chemical preconcentration have applied for multi-elemental analysis of Damascus drinking water. Water was taken directly from taps of several city sectors and analyzed for the following trace elements: Ti, V, Cr, Fe, Co, Ni, Cu, Zn, Se and Pb. The detection limits were found to be in the range of 0.1 to 0.4 μg/l. The mean levels of trace elements in the Damascus drinking water were below the World Health Organization drinking water quality guidelines. (author)

  8. Fast neutron activation analysis and radioisotope X-ray fluorescence study on KALEWA and NAMMA coal

    Energy Technology Data Exchange (ETDEWEB)

    Naing-Win, [Arts and Science University, Yangon (Myanmar)

    1981-07-01

    Kalewa coal was studied with Fast Neutron Activation Analysis (FNAA) technique, employing KAMAN A-710 neutron generator and HP(Ge) detector coupled to ``Canberra`` series 30 MCA. Sequential irradiation and dual aluminium foil monitoring method was employed. Simultaneous multielement analysis was carried out. Namma Coal was studied with radioisotope X-ray Fluorescence (XRF) technique, employing Co-57 exciter source and HP(Ge) detector coupled to ``Canberra`` series 40 MCA. In both FNAA and XRF study, the results obtained were compared to that obtained with Atomic Absorption Spectrophotometry (AAS) technique. Finally, the results were reviewed together with those obtained from similar work on coal with FNAA and XRF techniques. (author).

  9. Synchrotron total reflection X-ray fluorescence at BL-16 microfocus beamline of Indus-2

    Energy Technology Data Exchange (ETDEWEB)

    Tiwari, M. K., E-mail: mktiwari@rrcat.gov.in; Singh, A. K., E-mail: mktiwari@rrcat.gov.in; Das, Gangadhar, E-mail: mktiwari@rrcat.gov.in; Chowdhury, Anupam, E-mail: mktiwari@rrcat.gov.in; Lodha, G. S., E-mail: mktiwari@rrcat.gov.in [Indus Synchrotrons Utilisation Division, Raja Ramanna Centre for Advanced Technology, Indore-452013 (India)

    2014-04-24

    Determination of ultra trace elements is important in many disciplines both in basic and applied sciences. Numerous applications show their importance in medical science, environmental science, materials science, food processing and semiconductor industries and in maintaining the quality control of ultra pure chemicals and reagents. We report commissioning of a synchrotron based total reflection x-ray fluorescence (TXRF) facility on the BL-16 microfocus beamline of Indus-2. This paper describes the performance of the BL-16 TXRF spectrometer and the detailed description of its capabilities through examples of measured results.

  10. Trace elements in tobacco and tobacco smoke by x-ray fluorescence technique

    International Nuclear Information System (INIS)

    Mishra, U.C.; Shaikh, G.N.; Sadasivan, S.

    1986-01-01

    Trace elements in tobacco and tobacco smoke of a large number of commonly available brands of cigarettes were analyzed by energy dispersive x-ray fluorescence. This work supplements the data on the same samples gathered by INAA and reported earlier. Data on some toxic elements like Pb, Cu and Ni that could not be measured by INAA are presented. A number of chewing and snuff tobacco samples were also analyzed. The concentrations of Ca, K, Cl, Br, Cu, Fe, Ni, Pb, Rb, Sr, Ti and Zn in all these samples are presented and their relative hazards are discussed. (author)

  11. The determination, by x-ray-fluorescence spectrometry, of gold in activated charcoal

    International Nuclear Information System (INIS)

    Austen, C.E.

    1977-01-01

    A rapid method is described for the determination of gold in activated charcoal by X-ray-fluorescence spectrometry. Compensation for matrix effects is achieved by means of platinum that is added for use as an internal standard. Calibration is achieved by use of a series of synthetic standards that are made by the spiking of barren charcoal with gold and platinum. The limit of determination is about 8 p.p.m. of gold, and the relative standard deviation is 1,2 per cent at a concentration level of 2300 p.p.m

  12. Pattern recognition on X-ray fluorescence records from Copenhagen lake sediments using principal component analysis

    DEFF Research Database (Denmark)

    Schreiber, Norman; Garcia, Emanuel; Kroon, Aart

    2014-01-01

    Principle Component Analysis (PCA) was performed on chemical data of two sediment cores from an urban fresh-water lake in Copenhagen, Denmark. X-ray fluorescence (XRF) core scanning provided the underlying datasets on 13 variables (Si, K, Ca, Ti, Cr, Mn, Fe, Ni, Cu, Zn, Rb, Cd, Pb). Principle......, Fe, Rb) and characterized the content of minerogenic material in the sediment. In case of both cores, PC2 was a good descriptor emphasized as the contamination component. It showed strong linkages with heavy metals (Cu, Zn, Pb), disclosing changing heavy-metal contamination trends across different...

  13. The use of x-ray fluorescence in the detection of inorganic elements in medicines drugs

    International Nuclear Information System (INIS)

    Freitas, Tatiana Pereira da Silva de; Ricci Junior, Eduardo; Zucchi, Orgheda Luiza Araujo Domingues

    1997-01-01

    Based on the instrumental wavelength dispersive X-ray fluorescence technique of analysis (W D-X R F), the inorganic elements in four samples of commercial medicines drugs were analyzed. The samples were used without any chemical separation of the elements or chemical pre-treatment. Eleven elements could be identified, namely Si, P, S, Cl, K, Ca, Ti, Mn, Fe, Cu and Zn. However, Si, Ti and Mn have not been mentioned in the printed instructions for the use of the medicines drugs by the producers. (author)

  14. A Fundamental Parameter-Based Calibration Model for an Intrinsic Germanium X-Ray Fluorescence Spectrometer

    DEFF Research Database (Denmark)

    Christensen, Leif Højslet; Pind, Niels

    1982-01-01

    A matrix-independent fundamental parameter-based calibration model for an energy-dispersive X-ray fluorescence spectrometer has been developed. This model, which is part of a fundamental parameter approach quantification method, accounts for both the excitation and detection probability. For each...... secondary target a number of relative calibration constants are calculated on the basis of knowledge of the irradiation geometry, the detector specifications, and tabulated fundamental physical parameters. The absolute calibration of the spectrometer is performed by measuring one pure element standard per...

  15. Application of direct peak analysis to energy dispersive x-ray fluorescence spectra

    International Nuclear Information System (INIS)

    Nielson, K.K.

    1977-07-01

    A modified Covell method for direct peak analysis has been applied to energy dispersive x-ray fluorescence spectra. The method is background independent and is well-suited to computerized data reduction. It provides acceptable precision, minimizes errors from instrumental gain shift, and permits peak overlap correction. Peak overlap errors exhibit both positive and negative nodes as a function of peak separation distance, and are corrected using concentration ratios determined from thin, single-element standards. Peak precisions and overlaps are evaluated as a function of window width to aid in width selection. Least-square polynomial smoothing prior to peak analysis significantly improves peak area precisions without significantly affecting their accuracies

  16. X-ray fluorescence holography studies for a Cu{sub 3}Au crystal

    Energy Technology Data Exchange (ETDEWEB)

    Dąbrowski, K.M., E-mail: karol.dabrowski@uj.edu.pl; Dul, D.T.; Jaworska-Gołąb, T.; Rysz, J.; Korecki, P.

    2015-12-01

    In this work we show that performing a numerical correction for beam attenuation and indirect excitation allows one to fully restore element sensitivity in the three-dimensional reconstruction of the atomic structure. This is exemplified by a comparison of atomic images reconstructed from holograms measured for ordered and disordered phases of a Cu{sub 3}Au crystal that clearly show sensitivity to changes in occupancy of the atomic sites. Moreover, the numerical correction, which is based on quantitative methods of X-ray fluorescence spectroscopy, was extended to take into account the influence of a disturbed overlayer in the sample.

  17. Elimination of matrix effect in quantitative analysis of elements using x-ray fluorescence

    International Nuclear Information System (INIS)

    Sampaio, R.V.

    1973-07-01

    The emission-transmission method of Leroux and Mahmud, an experimental technique for compensating matrix effects in photon excited X-ray fluorescence analysis, was used to determine the concentration of lead and antimony in pellets of galalith. The effect of interfering elements was studied by adding various concentrations of mercury and tin to the respective pellets. To illustrate possible environmental applications, a number of pellets was prepared from leaves of almond trees located in different regions of Rio de Janeiro. Lead concentrations were determined for the dried leaf material and showed values ranging from 50 to 145 parts per million [pt

  18. Old traces, read anew - 'The Reading Hermit' painting in the light of X-ray fluorescence

    OpenAIRE

    Seim, C.; Laurenze-Landsberg, C.; Schröder-Smeibidl, B.; Mantouvalou, I.; Boer, C. de; Kanngießer, B.

    2014-01-01

    There exist several very similar looking versions of the painting ‘The Reading Hermit’, all allegedly painted by Rembrandt Harmenszoon van Rijn (approx. in ∼1630 A.D., Leiden). The classification of Rembrandt's paintings, which were produced by Rembrandt himself, in his academy by his students and the ones being mere copies is a crucial and difficult task. We gathered background evidence and performed elemental analyses by non-destructive micro-X-ray fluorescence (micro-XRF) in order to eluci...

  19. Separation of substandard tin ores by x-ray fluorescence method

    International Nuclear Information System (INIS)

    Kotler, N.I.; Konovalov, V.M.; Kamenskij, Yu.V.; Neverov, A.D.; Ogorodnikov, Yu.V.

    1987-01-01

    Analysis of pure tin ores on X-ray fluorescence separation (XFS) is carried out. The volumes of lump sampling are substantiated; several variants of technical and economical efficiency of XFS application have been calculated. It is shown that at XFS of -400+25 mm classes conditional as to tin content intermediate product with high efficiency factor may be prepared. Separation of -25+10 mm class is unsuitable, as it doesn't allow to increase tin content to conditional, and the process efficiency is low

  20. Computer simulation for synchrotron radiation based X-ray fluorescent microtomography

    International Nuclear Information System (INIS)

    Deng Biao; Yu Xiaohan; Xu Hongjie

    2007-01-01

    Synchrotron radiation based fluorescent microtomography (SR-XFMT) is a nondestructive technique for detecting elemental composition and distribution inside a specimen with high spatial resolution and sensitivity, and will be an optional experimental technique at SSRF hard X-ray micro-focusing beamline now under construction. In this paper, the principles and developments of SR-XFMT are briefly introduced. Computer simulation of SR-XFMT experiment is performed. The image of the simulated sample is reconstructed using Filtered Back Projection (FBP), Algebraic Reconstruction Techniques (ART) and modified FBP with absorption correction. The qualities of the reconstructed images are analyzed and compared. The validity of these reconstruction techniques is discussed. (authors)

  1. Analysis of zirconia and alumina products and raw materials by X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Boix, A.; Draignaud, M.; Debras-Guedon, J.

    1975-01-01

    An X-ray fluorescence spectrometric method of analysis for various raw materials and materials based on zirconia, alumina and silica was developed. The measurements were made on beads made from the sample and a borax based flux, in the proportions 1:10. A repeatability test allowed the relative standard deviation to be estimated for the three main constituents: SiO 2 , Al 2 O 3 , ZrO 2 : it is at the level of 1%. The element analyzed are: Si, Al, Fe, Ti, Ca, Hf and Zr [fr

  2. Exploratory analysis for the identification of false banknotes using portable X-ray Fluorescence spectrometer.

    Science.gov (United States)

    Zamalloa Jara, M A; Luízar Obregón, C; Araujo Del Castillo, C

    2018-05-01

    The aim of this study was to verify if a portable X-ray fluorescence (pXRF) spectrometer can recognize the security features in banknotes that are reproducible by counterfeiters. Peruvian Nuevo Sol banknotes were studied: 4 genuine and 3 fake ones, in 11 points of analysis for each one, at all 77 data set. The correlation analysis of spectra among original notes was 1.0, and there was no correlation with fake banknotes. pXRF prove that two security features were reproducible for counterfeiters. Copyright © 2018 The Authors. Published by Elsevier Ltd.. All rights reserved.

  3. Selection of polychlorinated plastics in plastic waste by X-ray fluorescence method

    International Nuclear Information System (INIS)

    Kumasaki, H.; Shinozaki, Y.

    1979-01-01

    The X-ray fluorescence method using a small source of 55 Fe was examined and found to be applicable for the selection of polychlorinated plastics from plastic waste in model areas in Tokyo designated for investigating their content in the waste. The weight ratios of soft and hard polychlorinated plastics to the total plastic waste estimated by this method were found to be 15.6% and 0.29% respectively. These values agree well with the results obtained with the Beilstein method. (author)

  4. The x-rays fluorescence applied to the analysis of alloys

    International Nuclear Information System (INIS)

    Gutierrez, D.A.

    1997-01-01

    This work is based on the utilization of the Fluorescence of X Rays. This technique of non destructive trial, has the purpose to establish a routine method, for the control of the conformation of industrial samples used. It makes an analysis with a combination of the algorithms of Rasberry-Heinrich and Claisse-Thinh. Besides, the numerical implementation of non usual techniques in this type of analysis. Such as the Linear Programming applied to the solution of super determined systems, of equations and the utilization of methods of relaxation to facilitate the convergence to the solutions. (author) [es

  5. X-ray fluorescence holography studies for a Cu3Au crystal

    Science.gov (United States)

    Dąbrowski, K. M.; Dul, D. T.; Jaworska-Gołąb, T.; Rysz, J.; Korecki, P.

    2015-12-01

    In this work we show that performing a numerical correction for beam attenuation and indirect excitation allows one to fully restore element sensitivity in the three-dimensional reconstruction of the atomic structure. This is exemplified by a comparison of atomic images reconstructed from holograms measured for ordered and disordered phases of a Cu3Au crystal that clearly show sensitivity to changes in occupancy of the atomic sites. Moreover, the numerical correction, which is based on quantitative methods of X-ray fluorescence spectroscopy, was extended to take into account the influence of a disturbed overlayer in the sample.

  6. Energy-dispersive X-ray fluorescence spectrometry of industrial paint samples

    International Nuclear Information System (INIS)

    Christensen, L.H.; Drabaek, I.

    1986-01-01

    An energy-dispersive X-ray fluorescence method for the direct, simultaneous determination of major and minor elements in coatings is described. The method relies on the back-scatter/fundamental parameter concept and provides a general solution to matrix problems. The method has been implemented and verified on spectrometers based both on tube excitation and radioisotope excitation. Results demonstrating some performance characteristics are presented. Sample inhomogeneity problems that impede quantification of low-Z elements in some types of paint are discussed. (Auth.)

  7. A high resolution x-ray fluorescence spectrometer for near edge absorption studies

    International Nuclear Information System (INIS)

    Stojanoff, V.; Hamalainen, K.; Siddons, D.P.; Hastings, J.B.; Berman, L.E.; Cramer, S.; Smith, G.

    1991-01-01

    A high resolution fluorescence spectrometer using a Johann geometry in a back scattering arrangement was developed. The spectrometer, with a resolution of 0.3 eV at 6.5 keV, combined with an incident beam, with a resolution of 0.7 eV, form the basis of a high resolution instrument for measuring x-ray absorption spectra. The advantages of the instrument are illustrated with the near edge absorption spectrum of dysprosium nitrate. 10 refs., 4 figs

  8. Determination of low concentrations of uranium in granite samples by X-ray fluorescence

    International Nuclear Information System (INIS)

    Roca, M.; Diaz-Guerra, J.P.

    1981-01-01

    An x-ray fluorescence method for the determination of uranium in granite samples for concentrations ranging from 3 to 100 ppm U 3 O 8 has been developed. To this purpose a sample holder, specially designed, allowing the irradiation of sample surfaces 42.5 mm in diameter and a molybdenum tube operating with a power of 2700 W (90 kV, 30 mA) are used. The background influence and the spectral interferences from rubidium and strontium have been taken into account and specific correction coefficients have been computed. A Basic program facilitates the report of the analytical results. (author)

  9. Thorium determination by X-ray Fluorescence Spectrometry in simulated thorex process solutions

    International Nuclear Information System (INIS)

    Yamaura, M.; Matsuda, H.T.

    1989-01-01

    The X-ray fluorescence method for thorium determination in aqueous and organic (TBP-n-dodecane) solutions is described. The thin film-technique for sample preparation and a suitable internal standard have been used. Some parameters as analytical line, internal standard, filter paper, paper geometry, sample volume and measurement conditions were studied. Uranium, fission products, corrosion products and thorex reagent components were studied as interfering elements in the thorium analysis, as well as the matrix effect by using the thorex process simulated solutions the method to thorium determination in irradiated thorium solutions was applied. (M.J.C.) [pt

  10. The analysis, by a fusion procedure and X-ray-fluorescence spectrometry, of silicates and slags

    International Nuclear Information System (INIS)

    Jacobs, J.J.; Balaes, A.M.E.

    1980-01-01

    A glass-disc fusion method is described for the analysis, by X-ray-fluorescence spectrometry, of slags and silicate materials. The data are corrected for detector dead time and short-term instrumental drift. Corrections are made for matrix variations by use of the Lachange-Traill mathematical model, and the results are processed on a mini-computer, an iterative procedure being used in the solving of the simultaneous equations. As the alpha-correction coefficients of the Lachange-Traill model are not truly constant, a modified version of the model is proposed [af

  11. Fluorescence X-ray microscopy on hydrated tributyltin-clay mineral suspensions

    Science.gov (United States)

    Neuhäusler, U.; Schmidt, C.; Hoch, M.; Susini, J.

    2003-03-01

    Using the scanning transmission X-ray microscope at ID21 beamline of the ESRF in fluorescence mode, we mapped tin at a bulk concentration of 1000 μg(Sn)/ml within hydrated tributyltin (TBT)-clay mineral (Kaolinite) dispersion with sub-300 nm spatial resolution. Using the L absorption edges of tin at 3929, 4156 and 4465 eV fluorescence radiation was excited in tin atoms with incident photon energies of 4 and 4.5 keV. When using 4 keV radiation, only tin fluorescence is excited. For 4.5 keV X rays, both the fluorescence of tin and calcium (which is present in the solid phase) can be measured. Methodologically, we were interested in assessing and proving the possibilities and limitations of fluorescence mapping using the L absorption edges of tin, where the fluorescence yield is significantly lower compared to other elements with their K edges in the same energy range. Scientifically, organotin-clay mineral interactions are of environmental concern because this factor influences significantly the distribution of toxic TBT in the aquatic System. On one hand, the half-life of TBT deposited to the sediment phase increases, and consequently the time of its bioavailability. On the other hand, the adsorption process is reversible, which means that contaminated sediments can act as a source of pollution. The adsorption and desorption effects can be studied directly with high spatial resolution and brought into connection to the surface properties of the clay mineral under study as well as to other experimental parameters, like pH or salinity.

  12. First combined total reflection X-ray fluorescence and grazing incidence X-ray absorption spectroscopy characterization of aeolian dust archived in Antarctica and Alpine deep ice cores

    Energy Technology Data Exchange (ETDEWEB)

    Cibin, G. [Diamond Light Source, Harwell Science and Innovation Campus, Didcot, Oxon OX110DE (United Kingdom); IMONT/EIM, Ente Italiano della Montagna, P.za dei Caprettari 70, 00176 Roma (Italy); Universita' degli Studi di Roma Tre, Dipartimento di Scienze Geologiche, L.go S. Leonardo Murialdo 1, 00146 Roma (Italy)], E-mail: giannantonio.cibin@diamond.ac.uk; Marcelli, A. [INFN - Laboratori Nazionali di Frascati, P.O. Box 13, 00044 Frascati (Roma) (Italy); Maggi, V. [Universita degli Studi di Milano-Bicocca, Dipartimento di Scienze dell' Ambiente e del Territorio, Piazza della Scienza 1, 20126 Milano (Italy); Sala, M. [Universita degli Studi di Milano-Bicocca, Dipartimento di Scienze dell' Ambiente e del Territorio, Piazza della Scienza 1, 20126 Milano (Italy); Universita degli Studi di Milano, Dipartimento di Scienze della Terra ' A. Desio' , Sez. Mineralogia, Via Mangiagalli 34, 20133 Milano (Italy); Marino, F.; Delmonte, B. [Universita degli Studi di Milano-Bicocca, Dipartimento di Scienze dell' Ambiente e del Territorio, Piazza della Scienza 1, 20126 Milano (Italy); Albani, S. [Universita degli Studi di Milano-Bicocca, Dipartimento di Scienze dell' Ambiente e del Territorio, Piazza della Scienza 1, 20126 Milano (Italy); Universita degli Studi di Siena, Dottorato in Scienze Polari, via Laterina 8, 53100 Siena (Italy); Pignotti, S. [IMONT/EIM, Ente Italiano della Montagna, P.za dei Caprettari 70, 00176 Roma (Italy)

    2008-12-15

    Aeolian mineral dust archived in polar and mid latitude ice cores represents a precious proxy for assessing environmental and climatic variations at different timescales. In this respect, the identification of dust mineralogy plays a key role. In this work we performed the first preliminary X-ray absorption spectroscopy (XAS) experiments on mineral dust particles extracted from Antarctic and from Alpine firn cores using grazing incidence geometry at the Fe K-edge. A dedicated high vacuum experimental chamber was set up for normal-incidence and total-reflection X-Ray Fluorescence and Absorption Spectroscopy analyses on minor amounts of mineral materials at the Stanford Synchrotron Radiation Laboratory. Results show that this experimental technique and protocol allows recognizing iron inclusion mineral fraction on insoluble dust in the 1-10 {mu}g range.

  13. Characterization of the Carancas-Puno meteorite by energy dispersive X-ray fluorescence, X-ray diffractometry and transmission Moessbauer spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Ceron Loayza, Maria L., E-mail: malucelo@hotmail.com; Bravo Cabrejos, Jorge A. [Universidad Nacional Mayor de San Marcos, Laboratorio de Analisis de Suelos, Facultad de Ciencias Fisicas (Peru)

    2011-11-15

    We report the results of the study of a meteorite that impacted an inhabited zone on 15 September 2007 in the neighborhood of the town of Carancas, Puno Region, about 1,300 km south of Lima. The analysis carried out by energy dispersive X-ray fluorescence, X-ray diffractometry and transmission Moessbauer spectroscopy (at room temperature and at 4.2 K), reveal the presence in the meteorite sample of magnetic sites assigned to taenite (Fe,Ni) and troilite (Fe,S) phases, and of two paramagnetic doublets assigned to Fe{sup 2 + }, one associated with olivine and the other to pyroxene. In accord with these results, this meteorite is classified as a type IV chondrite meteorite.

  14. A new detector system for low energy X-ray fluorescence coupled with soft X-ray microscopy: First tests and characterization

    Energy Technology Data Exchange (ETDEWEB)

    Gianoncelli, Alessandra, E-mail: alessandra.gianoncelli@elettra.eu [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Bufon, Jernej [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); University of Trieste, Piazzale Europa 1, Trieste 34127 (Italy); Ahangarianabhari, Mahdi [Politecnico di Milano, Via Anzani 42, Como 22100 (Italy); INFN Milano, Via Celoria 16, Milano 20133 (Italy); Altissimo, Matteo [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Bellutti, Pierluigi [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Bertuccio, Giuseppe [Politecnico di Milano, Via Anzani 42, Como 22100 (Italy); INFN Milano, Via Celoria 16, Milano 20133 (Italy); Borghes, Roberto [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Carrato, Sergio [University of Trieste, Piazzale Europa 1, Trieste 34127 (Italy); Cautero, Giuseppe [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Fabiani, Sergio [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Giacomini, Gabriele [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Giuressi, Dario [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Kourousias, George [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Menk, Ralf Hendrik [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Picciotto, Antonino; Piemonte, Claudio [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Rachevski, Alexandre [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); and others

    2016-04-21

    The last decades have witnessed substantial efforts in the development of several detector technologies for X-ray fluorescence (XRF) applications. In spite of the increasing trend towards performing, cost-effective and reliable XRF systems, detectors for soft X-ray spectroscopy still remain a challenge, requiring further study, engineering and customization in order to yield effective and efficient systems. In this paper we report on the development, first characterization and tests of a novel multielement detector system based on low leakage current silicon drift detectors (SDD) coupled to ultra low noise custom CMOS preamplifiers for synchrotron-based low energy XRF. This new system exhibits the potential for improving the count rate by at least an order of magnitude resulting in ten-fold shorter dwell time at an energy resolution similar to that of single element silicon drift detectors.

  15. First combined total reflection X-ray fluorescence and grazing incidence X-ray absorption spectroscopy characterization of aeolian dust archived in Antarctica and Alpine deep ice cores

    International Nuclear Information System (INIS)

    Cibin, G.; Marcelli, A.; Maggi, V.; Sala, M.; Marino, F.; Delmonte, B.; Albani, S.; Pignotti, S.

    2008-01-01

    Aeolian mineral dust archived in polar and mid latitude ice cores represents a precious proxy for assessing environmental and climatic variations at different timescales. In this respect, the identification of dust mineralogy plays a key role. In this work we performed the first preliminary X-ray absorption spectroscopy (XAS) experiments on mineral dust particles extracted from Antarctic and from Alpine firn cores using grazing incidence geometry at the Fe K-edge. A dedicated high vacuum experimental chamber was set up for normal-incidence and total-reflection X-Ray Fluorescence and Absorption Spectroscopy analyses on minor amounts of mineral materials at the Stanford Synchrotron Radiation Laboratory. Results show that this experimental technique and protocol allows recognizing iron inclusion mineral fraction on insoluble dust in the 1-10 μg range

  16. Elemental analysis of the ancient bronze coins by x-ray fluorescence technique using simultaneously radioisotope source and x-ray tube

    International Nuclear Information System (INIS)

    Nguyen The Quynh; Truong Thi An; Tran Duc Thiep; Nguyen Dinh Chien; Dao Tran Cao; Nguyen Quang Liem

    2004-01-01

    The results on elemental analysis of the Vietnamese ancient bronze coins during the time of the Nguyen dynasty (19th century) are presented. The samples were provided by the vietnam National Historical Museum and the elemental analysis was performed on the home-made model EDS-XT-99-01 X-ray fluorescence spectrometer in the Institute of Materials Science, NCST of Vietnam. The samples exited simultaneously by radioisotope source and X-ray tube. The analytical results show the similarity in the elemental composition of the coins issued by different kings of the Nguyen dynasty, but there is the difference in the concentration of the used elements. Another interesting point is that all the coins have zinc (Zn) in their composition, which shows clearly the influence of the occidental metallurgical technology on the money-making technique in Vietnam during the 19th century. (author)

  17. Mineralogical analysis of clays in hardsetting soil horizons, by X-ray fluorescence and X-ray diffraction using Rietveld method

    International Nuclear Information System (INIS)

    Prandel, L.V.; Saab, S.C.; Brinatti, A.M.; Giarola, N.F.B.; Leite, W.C.; Cassaro, F.A.M.

    2014-01-01

    Diffraction and spectroscopic techniques have been shown to be suitable for obtaining physical and mineralogical properties in polycrystalline soil samples, and also in their precursor compounds. For instance, the X-ray fluorescence (XRF) spectroscopy allows obtaining the elemental composition of an investigated sample, while the X-ray diffraction (XRD) technique permits obtaining qualitative and quantitative composition of the soil minerals through the Rietveld method (RM). In this study Yellow Latosol (Oxisol), Yellow Argisol (Ultisol) and Gray Argisol (Ultisol) soil samples, classified as “hardsetting soils”, extracted from areas located at Northeast and Southeast of Brazilian coast were investigated. The soils and their fractions were analyzed in an EDX-700 and an XRD-6000 (Cu K α radiation). XRF results indicate high percentages of Si and Al, and small percentage of Fe and Ti in the investigated samples. The DRX data and RM indicate that there was a predominance of kaolinite and halloysite minerals (kaolin group minerals) in the clay fractions, which are presumably responsible for the formation of kaolinitic plasma in these soils. Also, the obtained results showed that the XRF, XRD techniques and RM were very helpful for investigating the mineralogical composition of a hardsetting soil. - Highlights: ► Elemental composition of soil samples through X-Ray fluorescence. ► Mineralogical quantification through X-ray diffraction and Rietveld method. ► Oxisol and Ultisol, Brazil ‘Barreiras’ formation. ► High amounts of Si and Al oxides and low amounts of Fe and Ti oxides. ► Predominance of kaolinite in the clay fraction

  18. Combined phase and X-Ray fluorescence imaging at the sub-cellular level

    International Nuclear Information System (INIS)

    Kosior, Ewelina

    2013-01-01

    This work presents some recent developments in the field of hard X-ray imaging applied to biomedical research. As the discipline is evolving quickly, new questions appear and the list of needs becomes bigger. Some of them are dealt with in this manuscript. It has been shown that the ID22NI beamline of the ESRF can serve as a proper experimental setup to investigate diverse aspects of cellular research. Together with its high spatial resolution, high flux and high energy range the experimental setup provides bigger field of view, is less sensitive to radiation damages (while taking phase contrast images) and suits well chemical analysis with emphasis on endogenous metals (Zn, Fe, Mn) but also with a possibility for exogenous one's like these found in nanoparticles (Au, Pt, Ag) study. Two synchrotron-based imaging techniques, fluorescence and phase contrast imaging were used in this research project. They were correlated with each other on a number of biological cases, from bacteria E.coli to various cells (HEK 293, PC12, MRC5VA, red blood cells). The explorations made in the chapter 5 allowed preparation of more established and detailed analysis, described in the next chapter where both techniques, X-ray fluorescence and phase contrast imaging, were exploited in order to access absolute metal projected mass fraction in a whole cell. The final image presents for the first time true quantitative information at the sub-cellular level, not biased by the cell thickness. Thus for the first time a fluorescence map serves as a complete quantitative image of a cell without any risk of misinterpretation. Once both maps are divided by each other pixel by pixel (fluorescence map divided by the phase map) they present a complete and final result of the metal (Zn in this work) projected mass fraction in ppm of dry weight. For the purpose of this calculation the analysis was extended to calibration (non-biological) samples. Polystyrene spheres of a known diameter and known

  19. [Simultaneous determination of multiple elements in airborne particulate samples by X-ray fluorescence spectrometry].

    Science.gov (United States)

    Takada, T; Hitosugi, M; Kadowaki, T; Kudo, M

    1983-07-01

    An energy dispersive X-ray fluorescence spectrometer (EDX) has been applied to determine multielements in the workplace air. The standards for X-ray fluorescence analysis were prepared by the chelate precipitation method on polyvinyl chloride (PVC) membrane filter. And, the specimens were prepared to deposit various metal compounds of different chemical forms by the suspension method on PVC membrane filter, and they were determined with EDX and atomic absorption spectrometer (AAS). The results obtained were as follows. Though there is a difference by each element, an amount less than 3 microgram/cm2 per unit area makes it possible to undergo multielement analysis, that is, is has no influence on fine particle effect (particle size; under 5 microns). Then, effects of the X-ray intensity by different chemical forms are negligible. At the presence the neighboring element and other elements this technique showed greater precision by carrying out on corrective treatment, etc. The coefficient of variation of this technique was in the range of 2.5-6.5% at DDTC-Cu of 0.5-5.0 micrograms/cm2, with the limit of detection for As : 0.002 microgram/cm2, Zn : 0.003 microgram/cm2, Pb : 0.003 microgram/cm2, Cu : 0.004 microgram/cm2, Ni : 0.003 microgram/cm2, Fe : 0.005 microgram/cm2, Mn : 0.008 microgram/cm2, Cr : 0.013 microgram/cm2, respectively. Aerosols collected at the workplace were analyzed with EDX and AAS, and the obtained results showed good agreement with such regression line as y = 1.04 chi + 0.04, the coefficient of correlation being r = 0.995. From these results, this technique was found to be a very excellent method for monitoring of multielements in the workplace air.

  20. Portable x-ray fluorescence for the analysis of chromium in nail and nail clippings

    International Nuclear Information System (INIS)

    Fleming, David E.B.; Ware, Chris S.

    2017-01-01

    Assessment of chromium content in human nail or nail clippings could serve as an effective biomarker of chromium status. The feasibility of a new portable x-ray fluorescence (XRF) approach to chromium measurement was investigated through analysis of nail and nail clipping phantoms. Five measurements of 180 s (real time) duration were first performed on six whole nail phantoms having chromium concentrations of 0, 2, 5, 10, 15, and 20 µg/g. Using nail clippers, these phantoms were then converted to nail clippings, and assembled into different mass groups of 20, 40, 60, 80, and 100 mg for additional measurements. The amplitude of the chromium Kα characteristic x-ray energy peak was examined as a function of phantom concentration for all measurement conditions to create a series of calibration lines. The minimum detection limit (MDL) for chromium was also calculated for each case. The chromium MDL determined from the whole nail intact phantoms was 0.88±0.03 µg/g. For the clipping phantoms, the MDL ranged from 1.2 to 3.3 µg/g, depending on the mass group analyzed. For the 40 mg clipping group, the MDL was 1.2±0.1 µg/g, and higher mass collections did not improve upon this result. This MDL is comparable to chromium concentration levels seen in various studies involving human nail clippings. Further improvements to the portable XRF technique would be required to detect chromium levels expected from the lower end of a typical population. - Highlights: • Portable x-ray fluorescence (XRF) was explored as a technique to assess levels of chromium in human nails or nail clippings. • Results were found to depend on the mass of clipping sample provided. • Minimum detection limits for chromium were similar to concentration levels found in previous studies of human nail clippings.