Sample records for hand-held dental x-ray

  1. Leakage and scattered radiation from hand-held dental x-ray unit

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Eun Kyung [Dankook Univ. School of Dentistry, Seoul (Korea, Republic of)

    2007-06-15

    To compare the leakage and scattered radiation from hand-held dental X-ray unit with radiation from fixed dental X-ray unit. For evaluation we used one hand-held dental X-ray unit and Oramatic 558 (Trophy Radiologie, France), a fixed dental X-ray unit. Doses were measured with Unfors Multi-O-Meter 512L at the right and left hand levels of X-ray tube head part for the scattered and leakage radiation when human skull DXTTR {iota}{iota}{iota} was exposed to both dental X-ray units. And for the leakage radiation only, doses were measured at the immediately right, left, superior and posterior side of the tube head part when air was exposed. Exposure parameters of hand-held dental X-ray unit were 70 kVp, 3 mA , 0.1 second, and of fixed X-ray unit 70 kVp, 8 mA, 0.45 second. The mean dose at the hand level when human skull DXTTR {iota}{iota}{iota} was exposed with portable X-ray unit 6.39 {mu}Gy, and the mean dose with fixed X-ray unit 3.03 {mu}Gy (p<0.001). The mean dose at the immediate side of the tube head part when air was exposed with portable X-ray unit was 2.97 {mu}Gy and with fixed X-ray unit the mean dose was 0.68 {mu}Gy (p<0.01). The leakage and scattered radiation from hand-held dental radiography was greater than from fixed dental radiography.

  2. Leakage and scattered radiation from hand-held dental x-ray unit

    International Nuclear Information System (INIS)

    Kim, Eun Kyung

    2007-01-01

    To compare the leakage and scattered radiation from hand-held dental X-ray unit with radiation from fixed dental X-ray unit. For evaluation we used one hand-held dental X-ray unit and Oramatic 558 (Trophy Radiologie, France), a fixed dental X-ray unit. Doses were measured with Unfors Multi-O-Meter 512L at the right and left hand levels of X-ray tube head part for the scattered and leakage radiation when human skull DXTTR ΙΙΙ was exposed to both dental X-ray units. And for the leakage radiation only, doses were measured at the immediately right, left, superior and posterior side of the tube head part when air was exposed. Exposure parameters of hand-held dental X-ray unit were 70 kVp, 3 mA , 0.1 second, and of fixed X-ray unit 70 kVp, 8 mA, 0.45 second. The mean dose at the hand level when human skull DXTTR ΙΙΙ was exposed with portable X-ray unit 6.39 μGy, and the mean dose with fixed X-ray unit 3.03 μGy (p<0.001). The mean dose at the immediate side of the tube head part when air was exposed with portable X-ray unit was 2.97 μGy and with fixed X-ray unit the mean dose was 0.68 μGy (p<0.01). The leakage and scattered radiation from hand-held dental radiography was greater than from fixed dental radiography

  3. Position statement on use of hand-held portable dental X-ray equipment

    International Nuclear Information System (INIS)

    2014-06-01

    The position statement focuses on justification in the medical field, in particular on the use of hand-held portable dental x-ray equipment. It supplements another HERCA position paper, providing a general overview of the use of all hand-held portable X-ray equipment. Key Messages: - HERCA finds that the use of hand-held portable X-ray devices should be discouraged except in special circumstances. - As a general rule, these devices should only be used in scenarios where an intraoral radiograph is deemed necessary for a patient and the use of a fixed or semi-mobile x-ray unit is impractical, e.g.: - nursing homes, residential care facilities or homes for persons with disabilities; - forensic odontology, - military operations abroad without dental facilities

  4. Extra-oral dental radiography for disaster victims using a flat panel X-ray detector and a hand-held X-ray generator.

    Science.gov (United States)

    Ohtani, M; Oshima, T; Mimasaka, S

    2017-12-01

    Forensic odontologists commonly incise the skin for post-mortem dental examinations when it is difficult to open the victim's mouth. However, it is prohibited by law to incise dead bodies without permission in Japan. Therefore, we attempted using extra-oral dental radiography, using a digital X-ray equipment with rechargeable batteries, to overcome this restriction. A phantom was placed in the prone position on a table, and three plain dental radiographs were used per case: "lateral oblique radiographs" for left and right posterior teeth and a "contact radiograph" for anterior teeth were taken using a flat panel X-ray detector and a hand-held X-ray generator. The resolving power of the images was measured by a resolution test chart, and the scattered X-ray dose was measured using an ionization chamber-type survey meter. The resolving power of the flat panel X-ray detector was 3.0 lp/mm, which was less than that of intra-oral dental methods, but the three extra-oral plain dental radiographs provided the overall dental information from outside of the mouth, and this approach was less time-consuming. In addition, the higher dose of scattered X-rays was laterally distributed, but the dose per case was much less than that of intra-oral dental radiographs. Extra-oral plain dental radiography can be used for disaster victim identification by dental methods even when it is difficult to open the mouth. Portable and rechargeable devices, such as a flat panel X-ray detector and a hand-held X-ray generator, are convenient to bring and use anywhere, even at a disaster scene lacking electricity and water.

  5. Maintaining radiation exposures as low as reasonably achievable (ALARA) for dental personnel operating portable hand-held x-ray equipment.

    Science.gov (United States)

    McGiff, Thomas J; Danforth, Robert A; Herschaft, Edward E

    2012-08-01

    Clinical experience indicates that newly available portable hand-held x-ray units provide advantages compared to traditional fixed properly installed and operated x-ray units in dental radiography. However, concern that hand-held x-ray units produce higher operator doses than fixed x-ray units has caused regulatory agencies to mandate requirements for use of hand-held units that go beyond those recommended by the manufacturer and can discourage the use of this technology. To assess the need for additional requirements, a hand-held x-ray unit and a pair of manikins were used to measure the dose to a simulated operator under two conditions: exposures made according to the manufacturer's recommendations and exposures made according to manufacturer's recommendation except for the removal of the x-ray unit's protective backscatter shield. Dose to the simulated operator was determined using an array of personal dosimeters and a pair of pressurized ion chambers. The results indicate that the dose to an operator of this equipment will be less than 0.6 mSv y⁻¹ if the device is used according to the manufacturer's recommendations. This suggests that doses to properly trained operators of well-designed, hand-held dental x-ray units will be below 1.0 mSv y⁻¹ (2% of the annual occupational dose limit) even if additional no additional operational requirements are established by regulatory agencies. This level of annual dose is similar to those reported as typical dental personnel using fixed x-ray units and appears to satisfy the ALARA principal for this class of occupational exposures.

  6. Algorithms for a hand-held miniature x-ray fluorescence analytical instrument

    International Nuclear Information System (INIS)

    Elam, W.T.; Newman, D.; Ziemba, F.

    1998-01-01

    The purpose of this joint program was to provide technical assistance with the development of a Miniature X-ray Fluorescence (XRF) Analytical Instrument. This new XRF instrument is designed to overcome the weaknesses of spectrometers commercially available at the present time. Currently available XRF spectrometers (for a complete list see reference 1) convert spectral information to sample composition using the influence coefficients technique or the fundamental parameters method. They require either a standard sample with composition relatively close to the unknown or a detailed knowledge of the sample matrix. They also require a highly-trained operator and the results often depend on the capabilities of the operator. In addition, almost all existing field-portable, hand-held instruments use radioactive sources for excitation. Regulatory limits on such sources restrict them such that they can only provide relatively weak excitation. This limits all current hand-held XRF instruments to poor detection limits and/or long data collection times, in addition to the licensing requirements and disposal problems for radioactive sources. The new XRF instrument was developed jointly by Quantrad Sensor, Inc., the Naval Research Laboratory (NRL), and the Department of Energy (DOE). This report describes the analysis algorithms developed by NRL for the new instrument and the software which embodies them

  7. Panoramic Dental X-Ray

    Science.gov (United States)

    ... Physician Resources Professions Site Index A-Z Panoramic Dental X-ray Panoramic dental x-ray uses a very small dose of ... x-ray , is a two-dimensional (2-D) dental x-ray examination that captures the entire mouth ...

  8. X-ray film digitization using a personal computer and hand-held scanner: a simple technique for storing images

    International Nuclear Information System (INIS)

    Munoz-Nunez, C. F.; Lloret-Alcaniz, A.

    1998-01-01

    To develop a simple, low-cost technique for the digitization of X-ray films for personal use. A 66-MHz 486 PC with 8 MB of RAM, a Logitech ScanMan 256 hand-held scanner and a standard negatoscope with the power source converted to direct current. Although the system was originally designed for the digitization of mammographies, it has also been used with computed tomography, magnetic resonance, digital angiography and ultrasonographic images, as well as plain X-rays. After a minimal training period, the system digitized X-ray films easily and rapidly. Although the scanning values vary depending on the type of image to be digitized, an input spatial resolution of 200 dpi and a contrast resolution of 256 levels of gray are generally adequate. Of the storage formats tested, JPEG presented the best quality/image size ratio. A simple, low-cost technique has been developed for the digitization of X-ray films. This technique enables the storage of images in a digital format, thus facilitating their presentation and transmission. (Author) 9 refs

  9. User certification of hand-held x-ray tube based analytical fluorescent devices in a canadian context

    International Nuclear Information System (INIS)

    Maharaj, H.P.

    2005-01-01

    Safety education aims to reduce personal injury and improve well being. This health promotion principle is applied in the case of hand-held open beam x-ray tube based analytical x-ray devices. Such devices not only are light weight and portable, but also present high radiation exposure levels at the beam exit port and potentially can be used in a variety of industrial applications for determination of material composition. There is much potential for radiation risks to occur with resultant adverse effects if such devices are not used by trained individuals within controlled environments. A level of radiation safety knowledge and understanding of the device design, construction and performance characteristics appear warranted. To reduce radiation risks, user certification at a federal level was introduced in 2004 based on International Standards Organization 20807, since that standard comprises elements commensurate with risk reduction strategies. Within these contexts, a federally certified user is deemed to have acquired a level of safety knowledge and skills to facilitate safe use of the device. Certification, however, does not absolve the holder from obligations of compliance with applicable provincial, territorial or federal laws respecting device operation. The union of federal certification and applicable legislative mandated operational criteria reduces radiation risks overall. (author)

  10. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral dental X-ray apparatus for panoramic dental radiography is described in detail. It comprises an electron gun having an elongated tubular target carrier extending into the patient's mouth. The carrier supports an inclined target for direction of an X-ray pattern towards a film positioned externally of the patient's mouth. Image definition is improved by a focusing anode which focuses the electron beam into a sharp spot (0.05 to 0.10 mm diameter) on the target. The potential on the focusing anode is adjustable to vary the size of the spot. An X-ray transmitting ceramic (oxides of Be, Al and Si) window is positioned adjacent to the front face of the target. The electron beam can be magnetically deflected to change the X-ray beam direction. (author)

  11. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral dental X-ray apparatus for panoramic radiography is described in detail. It comprises a tubular target carrier supporting at its distal end a target with an inclined forward face. Image definition is improved by positioning in the path of the X-rays a window of X-ray transmitting ceramic material, e.g. 90% oxide of Be, or Al, 7% Si0 2 . The target carrier forms a probe which can be positioned in the patient's mouth. X-rays are directed forwardly and laterally of the target to an X-ray film positioned externally. The probe is provided with a detachable sleeve having V-form arms of X-ray opaque material which serve to depress the tongue out of the radiation path and also shield the roof of the mouth and other regions of the head from the X-ray pattern. A cylindrical lead shield defines the X-ray beam angle. (author)

  12. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral X-ray apparatus which reduces the number of exposures necessary to obtain panoramic dental radiographs is described in detail. It comprises an electron gun, a tubular target carrier projecting from the gun along the beam axis and carrying at its distal end a target surrounded by a shield of X-ray opaque material. This shield extends forward and laterally of the target and has surfaces which define a wedge or cone-shaped radiation pattern delimited vertically by the root tips of the patient's teeth. A film holder is located externally of the patient's mouth. A disposable member can fit on the target carrier to depress the patient's tongue out of the radiation pattern and to further shield the roof of the mouth. The electron beam can be magnetically deflected to change the X-ray beam direction. (author)

  13. X-ray exposures to dental patients

    International Nuclear Information System (INIS)

    McKlveen, J.W.

    1980-01-01

    An elastic mask worn by patients and a skeleton encased in plastic were instrumented with LiF thermoluminescent dosimeters to determine radiation exposures delivered from full-face diagnostic dental X-rays. Measurements were made using various panoramic radiographical and periapical machines. Locations of interest included skin surface, eyes, upper and lower teeth and thyroid. Exposures in the 100 mR range were common and a maximum of over 6000 mR was measured in the teeth region during a full-face examination with a periapical unit. In general, exposures received from periapical equipment were several times those obtained from panoramic devices. (author)

  14. Dental x-ray image segmentation

    Science.gov (United States)

    Said, Eyad; Fahmy, Gamal F.; Nassar, Diaa; Ammar, Hany

    2004-08-01

    Law enforcement agencies have been exploiting biometric identifiers for decades as key tools in forensic identification. With the evolution in information technology and the huge volume of cases that need to be investigated by forensic specialists, it has become important to automate forensic identification systems. While, ante mortem (AM) identification, that is identification prior to death, is usually possible through comparison of many biometric identifiers, postmortem (PM) identification, that is identification after death, is impossible using behavioral biometrics (e.g. speech, gait). Moreover, under severe circumstances, such as those encountered in mass disasters (e.g. airplane crashers) or if identification is being attempted more than a couple of weeks postmortem, under such circumstances, most physiological biometrics may not be employed for identification, because of the decay of soft tissues of the body to unidentifiable states. Therefore, a postmortem biometric identifier has to resist the early decay that affects body tissues. Because of their survivability and diversity, the best candidates for postmortem biometric identification are the dental features. In this paper we present an over view about an automated dental identification system for Missing and Unidentified Persons. This dental identification system can be used by both law enforcement and security agencies in both forensic and biometric identification. We will also present techniques for dental segmentation of X-ray images. These techniques address the problem of identifying each individual tooth and how the contours of each tooth are extracted.

  15. Panoramic dental X-ray machine X-motion drive

    International Nuclear Information System (INIS)

    Cushman, R.H.; Flynn, J.J.

    1980-01-01

    A panoramic dental x-ray machine is described which provides continuous and discontinuous radiographic images of the dental arch area of a patient. The systems for moving the chair and the column which carries the x-ray source and camera are specified. (U.K.)

  16. Electronic quality control on dental x-rays equipment

    International Nuclear Information System (INIS)

    Pomares C, Martin

    1996-09-01

    A brief description of dental x-ray equipment is done. The non-invasive quality control is treated as than the responsibilities from the dentists to the patient and the equipment. A propose for quality control for dental x-ray equipment, film and developer is include

  17. X-ray beam qualities for dental radiology purposes

    International Nuclear Information System (INIS)

    Santos, Marcus Aurelio P. dos; Fragoso, Maria da Conceicao de F.; Lima, Ricardo de A.; Hazim, Clovis A.

    2009-01-01

    In order to establish characteristics or properties of equipment for diagnostic radiology, e.g. ion chambers and semiconductor detectors, calibration laboratories offer a set of well-defined radiation conditions, called X-ray qualities, which can be used for many Physics studies and medical purposes. The standardization of radiation qualities has been carried out in several fields of study, but little attention has been given to the area of dental radiology, mainly for medical and physical applications using single-phase units with half-wave rectification. For this reason, a single-phase dental unit with adjustable peak voltage and tube current, called 'variable potential X-ray equipment', was developed aiming to define X-ray beam qualities for test and calibrations purposes. X-ray spectra at 50, 60 and 70 kVp were determined by using a CdTe detector and compared with those obtained for ten commercial X-ray dental units. As a result of this study, a set of X-ray qualities for the variable potential X-ray equipment was determined. The X-ray qualities spectra were utilized as reference for determination of a new set of X-ray qualities characterized for a constant potential X-ray equipment. Thus, sets of X-ray qualities were standardized and implemented in two X-ray laboratories: one with the variable potential X-ray equipment and other with constant potential X-ray equipment. These reference X-ray beam qualities should be used for test and calibration purposes involving scientific studies and services. (author)

  18. Use and regulatory control of dental X-ray installations

    International Nuclear Information System (INIS)

    1999-01-01

    In the guide the safety requirements concerning dental X-ray installations and their use, prerequisities for exemption from a safety licence, and regulatory control are presented. The guide applies to conventional dental X-ray installations, by which an image is created on an X-ray film or other image receptor placed inside the mouth, and panorama tomography installations for dentition and the cephalostats associated with these. The guide does not apply to multitechnique tomography installations intended for the special imaging of the skull or jaws

  19. Quality control procedure for dental x-ray film processing

    International Nuclear Information System (INIS)

    Tingey, D.R.C.

    1983-08-01

    Methods of obtaining the optimum processing from dental films are discussed. A method of quality control of developing conditions for dental x-ray films has been developed. It is relatively easy to maintain and is sufficiently accurate for practical purposes

  20. Digital Dental X-ray Database for Caries Screening

    Science.gov (United States)

    Rad, Abdolvahab Ehsani; Rahim, Mohd Shafry Mohd; Rehman, Amjad; Saba, Tanzila

    2016-06-01

    Standard database is the essential requirement to compare the performance of image analysis techniques. Hence the main issue in dental image analysis is the lack of available image database which is provided in this paper. Periapical dental X-ray images which are suitable for any analysis and approved by many dental experts are collected. This type of dental radiograph imaging is common and inexpensive, which is normally used for dental disease diagnosis and abnormalities detection. Database contains 120 various Periapical X-ray images from top to bottom jaw. Dental digital database is constructed to provide the source for researchers to use and compare the image analysis techniques and improve or manipulate the performance of each technique.

  1. Controlled Use of X-Ray in Dental Practice

    Directory of Open Access Journals (Sweden)

    Lesbia Rosa Tirado-Amado

    2015-01-01

    Full Text Available Abstract: Objective: To contribute in an informative and critical way to the promotion of controlled use of X-ray during dental consulting and dental care. Introduction: The practice of dentistry as a health care service requires the use of complementary tools for proper diagnosis, treatment and follow-up in patients with different clinical conditions. For that reason, the use of x-ray by dentists is common and very useful, but it involves radiation exposure to the patient and the dental professional or dental student. Although, often people do not pay attention to this exposition in regard to biological effects, because they are considered too low to generate significant biological effects. Conclusion: It is necessary to consider the risk of accumulative doses for constant exposure in dentist and dental students. Moreover, in patients, because a synergism with other radiations can occur, they can be exposed because of the attention in other areas of health. For these reasons, it is necessary to promote awareness and knowledge on basic aspects of the controlled use of X-ray in the dental care, being also aware of the need to strengthen the knowledge of the basics aspects that lead to decreased risk of biological effects from its proper use in the dental care.

  2. Radiation protection requirements for dental X-ray diagnostic facilities

    International Nuclear Information System (INIS)

    Taschner, P.; Koenig, W.; Andreas, M.; Trinius, W.

    1976-01-01

    On the basis of radiation protection regulations the planning of dental X-ray facilities is discussed considering organizational, technical and structural measures suitable for fulfilling protection requirements. Finally, instructions are given aimed at reducing radiation doses to personnel and patients. (author)

  3. Radiation protection requirements for dental X-ray diagnostic facilities

    Energy Technology Data Exchange (ETDEWEB)

    Taschner, P; Koenig, W [Staatliches Amt fuer Atomsicherheit und Strahlenschutz, Berlin (German Democratic Republic); Andreas, M [Karl-Marx-Universitaet, Leipzig (German Democratic Republic). Fachrichtung Stomatologie; Trinius, W [Karl-Marx-Universitaet, Leipzig (German Democratic Republic). Radiologische Klinik

    1976-03-01

    On the basis of radiation protection regulations the planning of dental X-ray facilities is discussed considering organizational, technical and structural measures suitable for fulfilling protection requirements. Finally, instructions are given aimed at reducing radiation doses to personnel and patients.

  4. Routine dental x-ray : a health hazards

    International Nuclear Information System (INIS)

    Parthasarathy, K.S.

    1978-01-01

    In orthopantography (OPG) technique, a single panoramic x-ray replaces a dozen or so conventional x-ray exposures required for a full-mouth examination. OPG, thus, reduces radiation dose to the patients. However, the very simplicity of the OPG technique may lead to its misuse. The dentists are tempted to take too many OPG films and thus exposing the patients unnecessarily to X radiation. Dentists are advised against making dental x-radiography a routine part of examinations. Greater care should be exercised particularly in the case of children while using OPG. (M.G.B.)

  5. Evaluation of the shielding of dental X-rays units

    International Nuclear Information System (INIS)

    Medrano, E.; Vega C, H. R.; Letechipia de L, C.; Hernandez D, V. M.; Salas L, M. A.

    2014-08-01

    The capacity of the walls of the dental radio-diagnostic rooms has been determined, to diminish the dose levels during the use of the X-rays equipment s. The study was carried out in the Dentistry Academic Unit of the campus Siglo X XI of the Universidad Autonoma de Zacatecas. The X-rays equipment s are a learning tool for the dentistry students and they are also used for offering health services to the population; for this reason is important to verify that the dose levels outside of the room walls are safe. During the evaluation process were used conservative approaches without prejudice of the thickness necessary in benefit of the radiological protection. Of the evaluation was found that all the walls satisfy their function thoroughly like barriers against the X-rays. (Author)

  6. Accidental over-exposure from dental X-ray equipment

    Energy Technology Data Exchange (ETDEWEB)

    Phillips, B G [National Radiological Protection Board, Harwell (UK)

    1976-07-01

    A description is given of an unusual dental X-ray procedure which resulted in accidental over-exposure both to the dentist and to several of his patients when a short-circuit was present in newly-installed equipment. The short-circuit by-passed the exposure control and energized the tube for certain orientations of the X-ray tube. The dentist left the patients, who wore protective aprons, to initiate the exposure themselves, using the control button. Although the warning lights were on, the dentist was not present in the room during the exposure, and the over-exposures were only detected when the developed X-ray films were found to be completely blackened. A reconstruction of the procedure enabled estimates to be made of the dose equivalents to the dentist's body and to the skin of the head, the eyes and the gonads of the patients. The dentist had overlooked several of basic principles recommended in the Code of Practice for the Protection of Persons against Ionizing Radiations from Medical and Dental Use (1972). It is pointed out that incidents involving failure of dental equipment (usually the timer mechanism) are not infrequent.

  7. Measurements of X ray absorbed doses to dental patients in two dental X ray units in Nigeria

    International Nuclear Information System (INIS)

    Ogundare, F.O.; Oni, O.M.; Balogun, F.A.

    2002-01-01

    Measurements of absorbed doses from radiographic examinations to various anatomical sites in the head and neck of patients with an average age of 45 years using intra-oral dental radiography have been carried out. LiF (TLD-100) dosemeters were used for the measurements of the absorbed dose. The measured absorbed doses to the various anatomical sites in the two units are reported, discussed and compared with results from the literature. Quality control measurements were also performed using a Victoreen quality control test device on the X ray units. The tube voltage accuracies for the two units were found to be within acceptable limits (less than ±10%). On the other hand the exposure time accuracies for these units have large deviations (>20%). These results and those that have been reported in the literature may be an indication that high patient doses are common in most dental X ray centres and countries. As a result of this, regular compliance and performance checks of dental diagnostic X ray equipment are essential in order to ensure proper performance and to minimise unnecessary patient and operator doses. (author)

  8. An Intraoral Miniature X-ray Tube Based on Carbon Nanotubes for Dental Radiography

    OpenAIRE

    Hyun Jin Kim; Hyun Nam Kim; Hamid Saeed Raza; Han Beom Park; Sung Oh Cho

    2016-01-01

    A miniature X-ray tube based on a carbon-nanotube electron emitter has been employed for the application to a dental radiography. The miniature X-ray tube has an outer diameter of 7 mm and a length of 47 mm. The miniature X-ray tube is operated in a negative high-voltage mode in which the X-ray target is electrically grounded. In addition, X-rays are generated only to the teeth directions using a collimator while X-rays generated to other directions are shielded. Hence, the X-ray tube can be ...

  9. Radiation doses for X-ray diagnosis teeth in dental medicine

    International Nuclear Information System (INIS)

    Direkov, Lyubomir

    2009-01-01

    X-rays are the first ionizing radiation, which are applied in medicine for diagnostic radiology and X-ray therapy. While in the beginning they are mainly used for X-ray photos of the chest /lungs and in severe fractures of the limbs, then in recent years they are widely applied in diagnostics of teeth in dental medicine. Considering that caries is a widespread disease, both in children and adults, and it requires repeated x-ray photographs of the damaged teeth for the individual, the total radiation doses, which reflect on people from the X-rays are at high values. In order to reduce external exposure to other organs /mainly thyroid gland/ by X-ray pictures of teeth, it should be used with special lead aprons with large coefficient of reduction. Keywords: doses of radiation, X-ray machines, dental, x-ray pictures of teeth, protection sources

  10. A miniature X-ray tube based on carbon nanotube for an intraoral dental radiography

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Hyun Jin; Park, Han Beom; Lee, Ju Hyuk; Cho, Sung Oh [KAIST, Daejeon (Korea, Republic of)

    2016-05-15

    The number of human teeth that can be radiographically taken is limited. Moreover, at least two X-ray shots are required to get images of teeth from both sides of the mouth. In order to overcome the disadvantages of conventional dental radiography, a dental radiograph has been proposed in which an X-ray tube is inserted into the mouth while an X-ray detector is placed outside the mouth. The miniature X-ray tube is required small size to insert into the mouth. Recently, we have fabricated a miniature x-ray tube with the diameter of 7 mm using a carbon nanotube (CNT) field. But, commercialized miniature X-ray tube were adopted a thermionic type using tungsten filament. The X-ray tubes adopted thermionic emission has a disadvantage of increasing temperature of x-ray tube. So it need to cooling system to cool x-ray tube. On the other hands, X-ray tubes adopted CNT field emitters don't need cooling systems because electrons are emitted from CNT by applying high voltage without heating. We have developed the miniature x-ray tube that produce x-ray with uniform spatial distribution based on carbon nanotube field emitters. The fabricated miniature x-ray tube can be stably and reliably operated at 50kV without any vacuum pump. The developed miniature X-ray tube was applied for intraoral dental radiography that employs an intra-oral CNT-based miniature X-ray tube and extra-oral X-ray detectors. An X-ray image of many teeth was successfully obtained by a single X-ray shot using the intra-oral miniature X-ray tube system. Furthermore, images of both molar teeth of pig were simultaneously obtained by a single X-ray shot. These results show that the intraoral dental radiography, which employs an intraoral miniature X-ray tube and an extraoral X-ray detector, performs better than conventional dental radiography.

  11. A miniature X-ray tube based on carbon nanotube for an intraoral dental radiography

    International Nuclear Information System (INIS)

    Kim, Hyun Jin; Park, Han Beom; Lee, Ju Hyuk; Cho, Sung Oh

    2016-01-01

    The number of human teeth that can be radiographically taken is limited. Moreover, at least two X-ray shots are required to get images of teeth from both sides of the mouth. In order to overcome the disadvantages of conventional dental radiography, a dental radiograph has been proposed in which an X-ray tube is inserted into the mouth while an X-ray detector is placed outside the mouth. The miniature X-ray tube is required small size to insert into the mouth. Recently, we have fabricated a miniature x-ray tube with the diameter of 7 mm using a carbon nanotube (CNT) field. But, commercialized miniature X-ray tube were adopted a thermionic type using tungsten filament. The X-ray tubes adopted thermionic emission has a disadvantage of increasing temperature of x-ray tube. So it need to cooling system to cool x-ray tube. On the other hands, X-ray tubes adopted CNT field emitters don't need cooling systems because electrons are emitted from CNT by applying high voltage without heating. We have developed the miniature x-ray tube that produce x-ray with uniform spatial distribution based on carbon nanotube field emitters. The fabricated miniature x-ray tube can be stably and reliably operated at 50kV without any vacuum pump. The developed miniature X-ray tube was applied for intraoral dental radiography that employs an intra-oral CNT-based miniature X-ray tube and extra-oral X-ray detectors. An X-ray image of many teeth was successfully obtained by a single X-ray shot using the intra-oral miniature X-ray tube system. Furthermore, images of both molar teeth of pig were simultaneously obtained by a single X-ray shot. These results show that the intraoral dental radiography, which employs an intraoral miniature X-ray tube and an extraoral X-ray detector, performs better than conventional dental radiography

  12. An Intraoral Miniature X-ray Tube Based on Carbon Nanotubes for Dental Radiography

    Directory of Open Access Journals (Sweden)

    Hyun Jin Kim

    2016-06-01

    Full Text Available A miniature X-ray tube based on a carbon-nanotube electron emitter has been employed for the application to a dental radiography. The miniature X-ray tube has an outer diameter of 7 mm and a length of 47 mm. The miniature X-ray tube is operated in a negative high-voltage mode in which the X-ray target is electrically grounded. In addition, X-rays are generated only to the teeth directions using a collimator while X-rays generated to other directions are shielded. Hence, the X-ray tube can be safely inserted into a human mouth. Using the intra-oral X-ray tube, a dental radiography is demonstrated where the positions of an X-ray source and a sensor are reversed compared with a conventional dental radiography system. X-ray images of five neighboring teeth are obtained and, furthermore, both left and right molar images are achieved by a single X-ray shot of the miniature X-ray tube.

  13. An intraoral miniature x-ray tube based on carbon nanotubes for dental radiography

    International Nuclear Information System (INIS)

    Kim, Hyun Jin; Kim, Hyun Nam; Raza, Hamid Saeed; Park, Han Beom; Cho, Sung Oh

    2016-01-01

    A miniature X-ray tube based on a carbon-nanotube electron emitter has been employed for the application to a dental radiography. The miniature X-ray tube has an outer diameter of 7 mm and a length of 47 mm. The miniature X-ray tube is operated in a negative high-voltage mode in which the X-ray target is electrically grounded. In addition, X-rays are generated only to the teeth directions using a collimator while X-rays generated to other directions are shielded. Hence, the X-ray tube can be safely inserted into a human mouth. Using the intra-oral X-ray tube, a dental radiography is demonstrated where the positions of an X-ray source and a sensor are reversed compared with a conventional dental radiography system. X-ray images of five neighboring teeth are obtained and, furthermore, both left and right molar images are achieved by a single X-ray shot of the miniature X-ray tube

  14. An intraoral miniature x-ray tube based on carbon nanotubes for dental radiography

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Hyun Jin; Kim, Hyun Nam; Raza, Hamid Saeed; Park, Han Beom; Cho, Sung Oh [Dept. of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon (Korea, Republic of)

    2016-06-15

    A miniature X-ray tube based on a carbon-nanotube electron emitter has been employed for the application to a dental radiography. The miniature X-ray tube has an outer diameter of 7 mm and a length of 47 mm. The miniature X-ray tube is operated in a negative high-voltage mode in which the X-ray target is electrically grounded. In addition, X-rays are generated only to the teeth directions using a collimator while X-rays generated to other directions are shielded. Hence, the X-ray tube can be safely inserted into a human mouth. Using the intra-oral X-ray tube, a dental radiography is demonstrated where the positions of an X-ray source and a sensor are reversed compared with a conventional dental radiography system. X-ray images of five neighboring teeth are obtained and, furthermore, both left and right molar images are achieved by a single X-ray shot of the miniature X-ray tube.

  15. 21 CFR 1000.60 - Recommendation on administratively required dental x-ray examinations.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Recommendation on administratively required dental x-ray examinations. 1000.60 Section 1000.60 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT... Recommendations § 1000.60 Recommendation on administratively required dental x-ray examinations. (a) The Food and...

  16. A Study of the Resolution of Dental Intraoral X-Ray Machines

    International Nuclear Information System (INIS)

    Kim, Seon Ju; Chung, Hyon De

    1990-01-01

    The purpose of this study was to assess the resolution and focal spot size of dental X-ray machines. Fifty dental X-ray machines were selected for measuring resolution and focal spot size. These machines were used in general dental clinics. The time on installation of the X-ray machine varies from 1 years to 10 years. The resolution of these machines was measured with the test pattern. The focal spot size of these machines was measured with the star test pattern. The following results were obtained: 1. The resolution of dental intraoral X-ray machines was not significantly changed in ten years. 2. The focal spot size of dental intraoral X-ray machines was not significantly increased in ten years. The statistical analysis between the mean focal spot size and nominal focal spot size was significant at the 0.05 level about the more than 3 years used machines.

  17. Exposure reduction in general dental practice using digital x-ray imaging system for intraoral radiography with additional x-ray beam filter

    International Nuclear Information System (INIS)

    Shibuya, Hitoshi; Mori, Toshimichi; Hayakawa, Yoshihiko; Kuroyanagi, Kinya; Ota, Yoshiko

    1997-01-01

    To measure exposure reduction in general dental practice using digital x-ray imaging systems for intraoral radiography with additional x-ray beam filter. Two digital x-ray imaging systems, Pana Digital (Pana-Heraus Dental) and CDR (Schick Technologies), were applied for intraoral radiography in general dental practice. Due to the high sensitivity to x-rays, additional x-ray beam filters for output reduction were used for examination. An Orex W II (Osada Electric Industry) x-ray generator was operated at 60 kVp, 7 mA. X-ray output (air-kerma; Gy) necessary for obtaining clinically acceptable images was measured at 0 to 20 cm in 5 cm steps from the cone tip using an ionizing chamber type 660 (Nuclear Associates) and compared with those for Ektaspeed Plus film (Eastman Kodak). The Pana Digital system was used with the optional filter supplied by Pana-Heraus Dental which reduced the output to 38%. The exposure necessary to obtain clinically acceptable images was only 40% of that for the film. The CDR system was used with the Dental X-ray Beam Filter Kit (Eastman Kodak) which reduced the x-ray output to 30%. The exposure necessary to obtain clinically acceptable images was only 20% of that for the film. The two digital x-ray imaging systems, Pana Digital and CDR, provided large dose savings (60-80%) compared with Ektaspeed Plus film when applied for intraoral radiography in general dental practice. (author)

  18. Dental x-ray validation study: comparison of information from patient interviews and dental charts

    International Nuclear Information System (INIS)

    Preston-Martin, S.; Bernstein, L.; Maldonado, A.A.; Henderson, B.E.; White, S.C.

    1985-01-01

    Information was collected from dentists of a subset of participants in a case-control interview study conducted in Los Angeles County, California, in August 1980-August 1981 to evaluate the relationship of dental x-rays to tumors of the parotid gland. Complete dental charts were available from 142 dentists of 84 cases and from 130 dentists of 79 control. Analysis of data from these interview chart comparisons indicates that recall appears to be unbiased since the measures of agreement between interview and dental chart data are similar for cases and controls. The authors further conclude that interview data alone may be used for case-control comparisons of dental x-ray exposure and would, because of unbiased misclassification, tend to underestimate the relative risks. 12 references, 3 tables

  19. High-contrast x-ray microtomography in dental research

    Science.gov (United States)

    Davis, Graham; Mills, David

    2017-09-01

    X-ray microtomography (XMT) is a well-established technique in dental research. The technique has been used extensively to explore the complex morphology of the root canal system, and to qualitatively and quantitatively evaluate root canal instrumentation and filling efficacy in extracted teeth; enabling different techniques to be compared. Densitometric information can be used to identify and map demineralized tissue resulting from tooth decay (caries) and, in extracted teeth, the method can be used to evaluate different methods of excavation. More recently, high contrast XMT is being used to investigate the relationship between external insults to teeth and the pulpal reaction. When such insults occur, fluid may flow through dentinal tubules as a result of cracking or porosity in enamel. Over time, there is an increase in mineralization along the paths of the tubules from the pulp to the damaged region in enamel and this can be visualized using high contrast XMT. The scanner used for this employs time-delay integration to minimize the effects of detector inhomogeneity in order to greatly increase the upper limit on signal-to-noise ratio that can be achieved with long exposure times. When enamel cracks are present in extracted teeth, the presence of these pathways indicates that the cracking occurred prior to extraction. At high contrast, growth lines are occasionally seen in deciduous teeth which may have resulted from periods of maternal illness. Various other anomalies in mineralization resulting from trauma or genetic abnormalities can also be investigated using this technique.

  20. Information about radiographic films rejects of dental x-rays

    International Nuclear Information System (INIS)

    Cezimbra, M.R.; Bernarsiuk, M.E.; Bauer V, E.

    1996-01-01

    The purpose of this research was to qualify and quantify the number of dental x-ray films rejected in a Porto Alegre clinic. As we analyzed the captured data, it was concluded that, our of 1066 peri-apical films, we had a total percentage of 4.5% in relation to the total of the exams made. This 4.5% consists of the following rejects: placement, patient movement, technical errors, diaphragm, too much clarity in the result, double exposure, prolongation, shortness, darkened for not have been shot, superposed film. Because of that, the rejection, due to the bad placement of the film, is the one with the larger percentage value, i.e., 1.22% of the 4.5%. With the knowledge of the types of rejects and their causes, it was possible to correct some sources of systematic errors minimizing the repetition of the exams, saving costs, time and diminishing the ionizing radiation exposure for the patient, odontologist and his technical staff, which will be proved. (authors). 4 refs., 1 tab

  1. Wavelet based Image Registration Technique for Matching Dental x-rays

    OpenAIRE

    P. Ramprasad; H. C. Nagaraj; M. K. Parasuram

    2008-01-01

    Image registration plays an important role in the diagnosis of dental pathologies such as dental caries, alveolar bone loss and periapical lesions etc. This paper presents a new wavelet based algorithm for registering noisy and poor contrast dental x-rays. Proposed algorithm has two stages. First stage is a preprocessing stage, removes the noise from the x-ray images. Gaussian filter has been used. Second stage is a geometric transformation stage. Proposed work uses two l...

  2. Lead foil in dental X-ray film: Backscattering rejection or image intensifier?

    International Nuclear Information System (INIS)

    Hönnicke, M.G.; Delben, G.J.; Godoi, W.C.; Swinka-Filho, V.

    2014-01-01

    Dental X-ray films are still largely used due to sterilization issues, simplicity and, mainly, economic reasons. These films almost always are double coated (double emulsion) and have a lead foil in contact with the film for X-ray backscattering rejection. Herein we explore the use of the lead foil as an image intensifier. In these studies, spatial resolution was investigated when images were acquired on the dental X-ray films with and without the lead foil. Also, the lead foil was subjected to atomic analysis (fluorescent measurements) and structure analysis (X-ray diffraction). We determined that the use of the lead foil reduces the exposure time, however, does not affect the spatial resolution on the acquired images. This suggests that the fluorescent radiation spread is smaller than the grain sizes of the dental X-ray films

  3. Lead foil in dental X-ray film: Backscattering rejection or image intensifier?

    Science.gov (United States)

    Hönnicke, M. G.; Delben, G. J.; Godoi, W. C.; Swinka-Filho, V.

    2014-11-01

    Dental X-ray films are still largely used due to sterilization issues, simplicity and, mainly, economic reasons. These films almost always are double coated (double emulsion) and have a lead foil in contact with the film for X-ray backscattering rejection. Herein we explore the use of the lead foil as an image intensifier. In these studies, spatial resolution was investigated when images were acquired on the dental X-ray films with and without the lead foil. Also, the lead foil was subjected to atomic analysis (fluorescent measurements) and structure analysis (X-ray diffraction). We determined that the use of the lead foil reduces the exposure time, however, does not affect the spatial resolution on the acquired images. This suggests that the fluorescent radiation spread is smaller than the grain sizes of the dental X-ray films.

  4. Terahertz pulsed imaging for the monitoring of dental caries: a comparison with x-ray imaging

    Science.gov (United States)

    Karagoz, Burcu; Kamburoglu, Kıvanc; Altan, Hakan

    2017-07-01

    Dental caries in sliced samples are investigated using terahertz pulsed imaging. Frequency domain terahertz response of these structures consistent with X-ray imaging results show the potential of this technique in the detection of early caries.

  5. Measurement of scattered and transmitted X-rays from intra-oral and panoramic dental X-ray equipment.

    Science.gov (United States)

    Holroyd, John Richard

    2018-04-10

    To quantify the levels of transmitted radiation arising from the use of intra-oral dental X-ray equipment and scattered radiation arising from the use of both intra-oral and panoramic X-ray equipment. Methods: Levels of scattered radiation were measured at 1 m from a phantom, using an 1800 cc ion chamber. Transmitted radiation was measured using both: i) a phantom and Dose Area Product (DAP) meter, ii) a patient and an 1800 cc ion chamber. Results: For intra-oral radiography the patient study gave a maximum transmission of 1.80% (range 0.04% to 1.80%, mean 0.26%) and the phantom study gave a maximum transmission of 6% (range 2% to 6%, mean 5%). The maximum scattered radiation, per unit DAP, was 5.5 nGy (mGy cm2)-1 at 70 kVp and a distance of 1 m. For panoramic radiography the maximum scattered radiation was 9.3 nGy (mGy cm2)-1 at 80 kVp and a distance of 1 m. Conclusions: Typical doses from scattered and transmitted radiation in modern dental practice have been measured and values are presented to enable the calculation of adequate protection measures for dental radiography rooms. Advances in knowledge: Previous studies have used a phantom and measured radiation doses at 1 m from the phantom to determine the radiation dose transmitted through a patient, whereas this study uses both patient and phantom measurements together with a large area dose meter, positioned to capture the entire X-ray beam, to ensure more realistic dose measurements can be made. © 2018 IOP Publishing Ltd.

  6. Lead aprons - should they be used for dental x-rays?

    International Nuclear Information System (INIS)

    Le Heron, J.

    1997-01-01

    When dental x-rays are performed it is now becoming increasingly common for dentists to offer their patients a lead apron to wear. However not all dentists are offering lead aprons. Are some dentists being negligent, and should the use of lead aprons for dental patients be mandatory? This article hopefully answers these questions, and in so doing presents a clear picture of what happens when the dentist presses the exposure button on the x-ray machine. (author). 1 tab

  7. Using computational modeling to compare X-ray tube Practical Peak Voltage for Dental Radiology

    International Nuclear Information System (INIS)

    Holanda Cassiano, Deisemar; Arruda Correa, Samanda Cristine; Monteiro de Souza, Edmilson; Silva, Ademir Xaxier da; Pereira Peixoto, José Guilherme; Tadeu Lopes, Ricardo

    2014-01-01

    The Practical Peak Voltage-PPV has been adopted to measure the voltage applied to an X-ray tube. The PPV was recommended by the IEC document and accepted and published in the TRS no. 457 code of practice. The PPV is defined and applied to all forms of waves and is related to the spectral distribution of X-rays and to the properties of the image. The calibration of X-rays tubes was performed using the MCNPX Monte Carlo code. An X-ray tube for Dental Radiology (operated from a single phase power supply) and an X-ray tube used as a reference (supplied from a constant potential power supply) were used in simulations across the energy range of interest of 40 kV to 100 kV. Results obtained indicated a linear relationship between the tubes involved. - Highlights: • Computational Model was developed to X-ray tube Practical Peak Voltage for Dental Radiology. • The calibration of X-rays tubes was performed using the MCNPX Monte Carlo code. • The energy range was 40–100 kV. • Results obtained indicated a linear relationship between the Dental Radiology and reference X-ray tubes

  8. Do Dental X-Rays Induce Genotoxicity and Cytotoxicity in Oral Mucosa Cells? A Critical Review.

    Science.gov (United States)

    Angelieri, Fernanda; Yujra, Veronica Quispe; Oshima, Celina Tizuko Fujiyama; Ribeiro, Daniel Araki

    2017-10-01

    Dental X-rays are widely used in clinical practice, since the technique is an important approach for diagnosing diseases in dental and periodontal tissues. However, it is widely known that radiation is capable of causing damage to cellular systems, such as genotoxicity or cytotoxicity. Thus, the aim of this review was to present a critical analysis regarding the studies published on genotoxicity and cytotoxicity induced by dental X-rays in oral mucosa cells. Such studies have revealed that some oral cell types are more sensitive than others following exposure to dental X-rays. Certainly, this review will contribute to a better understanding of this matter as well as to highlighting perspectives for further studies. Ultimately, such data will promote better safety for both patients and dental professionals. Copyright© 2017, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved.

  9. Dental x-rays and the risk of thyroid cancer: A case-control study

    International Nuclear Information System (INIS)

    Memon, Anjum; Godward, Sara; Williams, Dillwyn; Siddique, Iqbal; Al-Saleh, Khalid

    2010-01-01

    The thyroid gland is highly susceptible to radiation carcinogenesis and exposure to high-dose ionising radiation is the only established cause of thyroid cancer. Dental radiography, a common source of low-dose diagnostic radiation exposure in the general population, is often overlooked as a radiation hazard to the gland and may be associated with the risk of thyroid cancer. An increased risk of thyroid cancer has been reported in dentists, dental assistants, and x-ray workers; and exposure to dental x-rays has been associated with an increased risk of meningiomas and salivary tumours. Methods. To examine whether exposure to dental x-rays was associated with the risk of thyroid cancer, we conducted a population-based case-control interview study among 313 patients with thyroid cancer and a similar number of individually matched (year of birth ± three years, gender, nationality, district of residence) control subjects in Kuwait. Results. Conditional logistic regression analysis, adjusted for other upper-body x-rays, showed that exposure to dental x-rays was significantly associated with an increased risk of thyroid cancer (odds ratio = 2.1, 95% confidence interval: 1.4, 3.1) (p=0.001) with a dose-response pattern (p for trend <0.0001). The association did not vary appreciably by age, gender, nationality, level of education, or parity. Discussion. These findings, based on self-report by cases/controls, provide some support to the hypothesis that exposure to dental x-rays, particularly multiple exposures, may be associated with an increased risk of thyroid cancer; and warrant further study in settings where historical dental x-ray records may be available.

  10. Dental x-rays and the risk of thyroid cancer: A case-control study

    Energy Technology Data Exchange (ETDEWEB)

    Memon, Anjum (Div. of Primary Care and Public Health, Brighton and Sussex Medical School (United Kingdom)), E-mail: a.memon@bsms.ac.uk; Godward, Sara (Dept. of Public Health and Primary Care, Univ. of Cambridge (United Kingdom)); Williams, Dillwyn (Thyroid Carcinogenesis Research Group, Strangeways Research Laboratories, Univ. of Cambridge (United Kingdom)); Siddique, Iqbal (Dept. of Medicine, Faculty of Medicine, Kuwait Univ. (Kuwait)); Al-Saleh, Khalid (Kuwait Cancer Control Centre, Ministry of Health (Kuwait))

    2010-05-15

    The thyroid gland is highly susceptible to radiation carcinogenesis and exposure to high-dose ionising radiation is the only established cause of thyroid cancer. Dental radiography, a common source of low-dose diagnostic radiation exposure in the general population, is often overlooked as a radiation hazard to the gland and may be associated with the risk of thyroid cancer. An increased risk of thyroid cancer has been reported in dentists, dental assistants, and x-ray workers; and exposure to dental x-rays has been associated with an increased risk of meningiomas and salivary tumours. Methods. To examine whether exposure to dental x-rays was associated with the risk of thyroid cancer, we conducted a population-based case-control interview study among 313 patients with thyroid cancer and a similar number of individually matched (year of birth +- three years, gender, nationality, district of residence) control subjects in Kuwait. Results. Conditional logistic regression analysis, adjusted for other upper-body x-rays, showed that exposure to dental x-rays was significantly associated with an increased risk of thyroid cancer (odds ratio = 2.1, 95% confidence interval: 1.4, 3.1) (p=0.001) with a dose-response pattern (p for trend <0.0001). The association did not vary appreciably by age, gender, nationality, level of education, or parity. Discussion. These findings, based on self-report by cases/controls, provide some support to the hypothesis that exposure to dental x-rays, particularly multiple exposures, may be associated with an increased risk of thyroid cancer; and warrant further study in settings where historical dental x-ray records may be available.

  11. Quality control for dental X-rays equipment and film developers

    International Nuclear Information System (INIS)

    Pomares C, Martin

    1998-01-01

    Quality control in five dental X-ray equipments, film and film developers was done. It was for evaluating the radiologic practices in the odontologic services too. This work was made based on international standards, the results will be used for future works in quality assurance in dental radiology

  12. Health risk assessment of doses to patients\\' eyes from dental X-ray ...

    African Journals Online (AJOL)

    The skin entry dose to patients\\' eyes during dental x-ray examination was carried out on one hundred and ten patients comprising infants and adult of both sexes. The dose measurements was performed at Alpha dental centre, Ibadan, using Lithium fluoride thermoluminiscent dosimeters (TLD). The results of the study ...

  13. X-Ray pictures of the developmental anomalis of the hard dental tissue

    International Nuclear Information System (INIS)

    Cecetkova, A.; Ondrasovicova, J.

    2008-01-01

    Dental anomalies are rare lessions of the hard dental tissue. They are as symptoms varies of the syndromes. They are as follow: hyperdoncia, hypodoncia, oligodoncia, anodoncia, mesiodens, macrodoncia and microdoncia. All of the above anomalis are detected by X-ray diagnostics ( intraoral and extra-oral radiography ). (authors)

  14. X-ray pickup device for dental radiography

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    The X-ray tube has got a tubular extension within which a W-shaped target is arranged and which can be inserted into the mouth opening of the patient. The target is inclined with respect to the axis of the beam of electrons striking it in such a way, that the X-radiation emitted by it is received either by the upper or the lower row of teeth, including the root area. A screen mounted on the extension causes only these semicircular areas to be irradiated. An X-ray film holder is also fastened to the extension and covers mouth and cheeks of the patient from outside. The focal spot on the target is adjustable so that a sharp picture can be produced. (RW) [de

  15. Dental X-ray diagnostic facility with means to choose the time of exposure

    Energy Technology Data Exchange (ETDEWEB)

    Krause, H; Grassme, U

    1978-02-16

    Equipment for dental X-ray diagnostics with automatic exposure control. In the patient's mouth there is arranged a film carrier containing in a pocket a dental X-ray film. A detector-transmitter assembly touches the pocket. It reacts to the density of the film and wireless transmits a signal to an antenna. This antenna is connected to a receiver shutting off the high voltage of the X-ray tube by means of a circuit if a desired dose value is reached. The detector-transmitter unit is an integrated component. It has got a luminescent layer with a light-sensitive detector or a detector directly sensitive to X-rays.

  16. Cephalometric landmark detection in dental x-ray images using convolutional neural networks

    Science.gov (United States)

    Lee, Hansang; Park, Minseok; Kim, Junmo

    2017-03-01

    In dental X-ray images, an accurate detection of cephalometric landmarks plays an important role in clinical diagnosis, treatment and surgical decisions for dental problems. In this work, we propose an end-to-end deep learning system for cephalometric landmark detection in dental X-ray images, using convolutional neural networks (CNN). For detecting 19 cephalometric landmarks in dental X-ray images, we develop a detection system using CNN-based coordinate-wise regression systems. By viewing x- and y-coordinates of all landmarks as 38 independent variables, multiple CNN-based regression systems are constructed to predict the coordinate variables from input X-ray images. First, each coordinate variable is normalized by the length of either height or width of an image. For each normalized coordinate variable, a CNN-based regression system is trained on training images and corresponding coordinate variable, which is a variable to be regressed. We train 38 regression systems with the same CNN structure on coordinate variables, respectively. Finally, we compute 38 coordinate variables with these trained systems from unseen images and extract 19 landmarks by pairing the regressed coordinates. In experiments, the public database from the Grand Challenges in Dental X-ray Image Analysis in ISBI 2015 was used and the proposed system showed promising performance by successfully locating the cephalometric landmarks within considerable margins from the ground truths.

  17. The reduction methods of operator's radiation dose for portable dental X-ray machines.

    Science.gov (United States)

    Cho, Jeong-Yeon; Han, Won-Jeong

    2012-08-01

    This study was aimed to investigate the methods to reduce operator's radiation dose when taking intraoral radiographs with portable dental X-ray machines. Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used. Operator's radiation dose was measured with an 1,800 cc ionization chamber (RadCal Corp.) at the hand level of X-ray tubehead and at the operator's chest and waist levels with and without the backscatter shield. The operator's radiation dose at the hand level was measured with and without lead gloves and with long and short cones. The backscatter shield reduced operator's radiation dose at the hand level of X-ray tubehead to 23 - 32%, the lead gloves to 26 - 31%, and long cone to 48 - 52%. And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%. When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

  18. Settling time of dental x-ray tube head after positioning

    International Nuclear Information System (INIS)

    Yun, Suk Ja; Kang, Byung Cheol; Wang, Se Myung; Koh, Chang Sung

    2002-01-01

    The aim of this study was to introduce a method of obtaining the oscillation graphs of the dental x-ray tube heads relative to time using an accelerometer. An Accelerometer, Piezotron type 8704B25 (Kistler Instrument Co., Amherst, NY, USA) was utilized to measure the horizontal oscillation of the x-ray tube head immediately after positioning the tube head for an intraoral radiograph. The signal from the sensor was transferred to a dynamic signal analyzer, which displayed the magnitude of the acceleration on the Y-axis and time lapse on the X-axis. The horizontal oscillation of the tube head was measured relative to time, and the settling time was also determined on the basis of the acceleration graphs for 6 wall type, 5 floor-fixed type, and 4 mobile type dental x-ray machines. The oscillation graphs showed that tube head movement decreased rapidly over time. The settling time varied with x-ray machine types. Wall-type x-ray machines had a settling time of up to 6 seconds, 5 seconds for fixed floor-types, and 11 seconds for the mobile-types. Using an accelerometer, we obtained the oscillation graphs of the dental x-ray tube head relative to time. The oscillation graph with time can guide the operator to decide upon the optimum exposure moment after xray tube head positioning for better radiographic resolution.

  19. Settling time of dental x-ray tube head after positioning

    Energy Technology Data Exchange (ETDEWEB)

    Yun, Suk Ja; Kang, Byung Cheol [Department of Oral and Maxillofacial Radiology, Chonnam National University, Gwangju (Korea, Republic of); Wang, Se Myung; Koh, Chang Sung [Department of Mechatronics, Kwangju Institute of Science and Technology, Gwangju (Korea, Republic of)

    2002-09-15

    The aim of this study was to introduce a method of obtaining the oscillation graphs of the dental x-ray tube heads relative to time using an accelerometer. An Accelerometer, Piezotron type 8704B25 (Kistler Instrument Co., Amherst, NY, USA) was utilized to measure the horizontal oscillation of the x-ray tube head immediately after positioning the tube head for an intraoral radiograph. The signal from the sensor was transferred to a dynamic signal analyzer, which displayed the magnitude of the acceleration on the Y-axis and time lapse on the X-axis. The horizontal oscillation of the tube head was measured relative to time, and the settling time was also determined on the basis of the acceleration graphs for 6 wall type, 5 floor-fixed type, and 4 mobile type dental x-ray machines. The oscillation graphs showed that tube head movement decreased rapidly over time. The settling time varied with x-ray machine types. Wall-type x-ray machines had a settling time of up to 6 seconds, 5 seconds for fixed floor-types, and 11 seconds for the mobile-types. Using an accelerometer, we obtained the oscillation graphs of the dental x-ray tube head relative to time. The oscillation graph with time can guide the operator to decide upon the optimum exposure moment after xray tube head positioning for better radiographic resolution.

  20. X-ray pick-up device for dental radiography

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    The X-ray tube has got a tabular extension at the front end of which a W-shaped target is mounted. The target hit by electrons emits X-radiation passing through a ceramic window in the surface area. The extension can be inserted into the mouth opening of the patient. For holding down the tongue and for shielding purposes it has got a device shaped like a sleeve with diverging wings fastened to it. Thex leave the ceramic window uncovered. (orig.) [de

  1. Enhancement of dental x-ray images by two channel image processing

    International Nuclear Information System (INIS)

    Mitra, S.; Yu, T.H.

    1991-01-01

    In this paper, the authors develop a new algorithm for the enhancement of low-contrast details of dental X-ray images using a two channel structure. The algorithm first decomposes an input image in the frequency domain into two parts by filtering: one containing the low frequency components and the other containing the high frequency components. Then these parts are enhanced separately using a transform magnitude modifier. Finally a contrast enhanced image is formed by combining these two processed pats. The performance of the proposed algorithm is illustrated through enhancement of dental X-ray images. The algorithm can be easily implemented on a personal computer

  2. A comparison of Kodak Ultraspeed and Ektaspeed plus dental X-ray films for the detection of dental caries.

    Science.gov (United States)

    Wong, A; Monsour, P A; Moule, A J; Basford, K E

    2002-03-01

    Using the fastest dental X-ray film available is an easy way of reducing exposure to ionizing radiation. However, the diagnostic ability of fast films for the detection of proximal surface caries must be demonstrated before these films will become universally accepted. Extracted premolar and molar teeth were arranged to simulate a bitewing examination and radiographed using Ultraspeed and Ektaspeed Plus dental X-ray films. Three different exposure times were used for each film type. Six general dentists were used to determine the presence and depth of the decay in the proximal surfaces of the teeth radiographed. The actual extent of the decay in the teeth was determined by sectioning the teeth and examining them under a microscope. There was no significant difference between the two films for the mean correct diagnosis. However, there was a significant difference between the means for the three exposure times used for Ultraspeed film. The practitioners used were not consistent in their ability to make a correct diagnosis, or for the film for which they got the highest correct diagnosis. Ektaspeed Plus dental X-ray film is just as reliable as Ultraspeed dental X-ray film for the detection of proximal surface decay. The effect of underexposure was significant for Ultraspeed, but not for Ektaspeed Plus. Patient exposure can be reduced significantly with no loss of diagnostic ability by changing from Ultraspeed X-ray film to Ektaspeed Plus X-ray film.

  3. Image quality assessment and medical physics evaluation of different portable dental X-ray units.

    Science.gov (United States)

    Pittayapat, Pisha; Oliveira-Santos, Christiano; Thevissen, Patrick; Michielsen, Koen; Bergans, Niki; Willems, Guy; Debruyckere, Deborah; Jacobs, Reinhilde

    2010-09-10

    Recently developed portable dental X-ray units increase the mobility of the forensic odontologists and allow more efficient X-ray work in a disaster field, especially when used in combination with digital sensors. This type of machines might also have potential for application in remote areas, military and humanitarian missions, dental care of patients with mobility limitation, as well as imaging in operating rooms. To evaluate radiographic image quality acquired by three portable X-ray devices in combination with four image receptors and to evaluate their medical physics parameters. Images of five samples consisting of four teeth and one formalin-fixed mandible were acquired by one conventional wall-mounted X-ray unit, MinRay 60/70 kVp, used as a clinical standard, and three portable dental X-ray devices: AnyRay 60 kVp, Nomad 60 kVp and Rextar 70 kVp, in combination with a phosphor image plate (PSP), a CCD, or a CMOS sensor. Three observers evaluated images for standard image quality besides forensic diagnostic quality on a 4-point rating scale. Furthermore, all machines underwent tests for occupational as well as patient dosimetry. Statistical analysis showed good quality imaging for all system, with the combination of Nomad and PSP yielding the best score. A significant difference in image quality between the combination of the four X-ray devices and four sensors was established (p1m: Rextar <0.2 microGy, MinRay <0.1 microGy). The present study demonstrated the feasibility of three portable X-ray systems to be used for specific indications, based on acceptable image quality and sufficient accuracy of the machines and following the standard guidelines for radiation hygiene. Copyright 2010 Elsevier Ireland Ltd. All rights reserved.

  4. Development of Object Simulator for Radiation Field of Dental X-Rays

    International Nuclear Information System (INIS)

    Silva, L F; Ferreira, F C L; Sousa, F F; Cardoso, L X; Vasconcelos, E D S; Brasil, L M

    2013-01-01

    In dentistry radiography is of fundamental importance to the dentist can make an accurate diagnosis. For this it is necessary to pay attention to the radiological protection of both the professional and the patient and control image quality for an accurate diagnosis. In this work, quality control tests were performed on X-ray machines in private dental intraoral in the municipality of Marabá, where they measured the diameters of the radiation field to see if these machines are in accordance with the recommendations, thus preventing the patient is exposed to a radiation field higher than necessary. We will study the results of each X-ray machine evaluated. For this we created a phantom to assess the size of the radiation field of X-ray dental, where we measure the radiation field of each device to see if they are in accordance with the recommendations of the ordinance No. 453/98 – MS

  5. Drywall plates evaluation as protection barriers in dental X-rays and mammography facilities

    International Nuclear Information System (INIS)

    Guevara R, V. Y.; Romero C, N.; Berrocal T, M.

    2014-08-01

    In the dental X-rays and mammography facilities, usually lead shielded walls as protective barriers are used. Lead is a good attenuator for X-rays, but has toxic properties and its cost is high. Mammography equipment s emit low-energy photons in the range of 25 keV to 35 keV, on current computers; the primary radiation beam is intercepted by the image receptor. Peri apical dental equipment s emit photons in the range of 50 to 90 keV, their filtration is fixed. These devices emit a collimated beam whose diameter is slightly larger than the diagonal dimension of a standard film of dental X-rays. When a dental x-ray is performed, the radiation beam is partially attenuated by the patient. Drywall is a material consisting of plasterboard between two cardboard layers, being its components gypsum and cellulose generally. It is used in construction for execution of interior walls, ceilings and wall coverings, could also serve as a replacement for lead as well as other materials. In this paper three drywall prototypes (Giplac), formed with 02, 04 and 06 drywall layers (13, 16 and 20 cm of thickness respectively) were tested as barriers against primary and secondary X-ray radiation that come from dental and mammography equipment s. The results show that the drywall prototype, 02 layers, efficiently attenuates the secondary radiation beam produced by conventional mammography equipment. And the prototype 04 and 06 layers, efficiently attenuates the primary radiation beam produced by peri apical dental equipment. (author)

  6. Hand-held medical robots.

    Science.gov (United States)

    Payne, Christopher J; Yang, Guang-Zhong

    2014-08-01

    Medical robots have evolved from autonomous systems to tele-operated platforms and mechanically-grounded, cooperatively-controlled robots. Whilst these approaches have seen both commercial and clinical success, uptake of these robots remains moderate because of their high cost, large physical footprint and long setup times. More recently, researchers have moved toward developing hand-held robots that are completely ungrounded and manipulated by surgeons in free space, in a similar manner to how conventional instruments are handled. These devices provide specific functions that assist the surgeon in accomplishing tasks that are otherwise challenging with manual manipulation. Hand-held robots have the advantages of being compact and easily integrated into the normal surgical workflow since there is typically little or no setup time. Hand-held devices can also have a significantly reduced cost to healthcare providers as they do not necessitate the complex, multi degree-of-freedom linkages that grounded robots require. However, the development of such devices is faced with many technical challenges, including miniaturization, cost and sterility, control stability, inertial and gravity compensation and robust instrument tracking. This review presents the emerging technical trends in hand-held medical robots and future development opportunities for promoting their wider clinical uptake.

  7. First do no harm - The impact of financial incentives on dental X-rays

    NARCIS (Netherlands)

    Chalkley, M.; Listl, S.

    2018-01-01

    This article assesses the impact of dentist remuneration on the incidence of potentially harmful dental X-rays. We use unique panel data which provide details of 1.3 million treatment claims by Scottish NHS dentists made between 1998 and 2007. Controlling for unobserved heterogeneity of both

  8. [Evaluation of dental X-ray apparatus in terms of patient exposure to ionizing radiation].

    Science.gov (United States)

    Olszewski, Jerzy; Wrzesień, Małgorzata

    2017-06-27

    The use of X-ray in dental procedures causes exposure of the patient to ionizing radiation. This exposure depends primarily on the parameters used in tooth examination. The aim of the study was to determine the patients exposure and to assess the technical condition of X-ray tubes. Seventeen hundred dental offices were covered by the questionnaire survey and 740 questionnaires were sent back. Direct measurements were performed in 100 units by using the thermoluminescent detectors and X-ray films. The results showed that the most commonly used exposure time is 0.22±0.16 s. The average entrance dose for the parameters used most commonly by dentists is 1.7±1.4 mGy. The average efficiency of X-ray tube estimated on the basis of exposures is 46.5±23.7 μGy/mAs. The study results indicate that the vast majority of X-ray tubes meet the requirements specified in the binding regulations. Med Pr 2017;67(4):491-496. This work is available in Open Access model and licensed under a CC BY-NC 3.0 PL license.

  9. Theory of X-ray microcomputed tomography in dental research: application for the caries research

    Directory of Open Access Journals (Sweden)

    Young-Seok Park

    2011-03-01

    Full Text Available Caries remains prevalent throughout modern society and is the main disease in the field of dentistry. Although studies of this disease have used diverse methodology, recently, X-ray microtomography has gained popularity as a non-destructive, 3-dimensional (3D analytical technique, and has several advantages over the conventional methods. According to X-ray source, it is classified as monochromatic or polychromatic with the latter being more widely used due to the high cost of the monochromatic source despite some advantages. The determination of mineral density profiles based on changes in X-ray attenuation is the principle of this method and calibration and image processing procedures are needed for the better image and reproducible measurements. Using this tool, 3D reconstruction is also possible and it enables to visualize the internal structures of dental caries. With the advances in the computer technology, more diverse applications are being studied, such automated caries assessment algorithms.

  10. Frequency of medical and dental x-ray examinations in the UK. 1997/98

    International Nuclear Information System (INIS)

    Tanner, R.; Wall, B.; Shrimpton, P.

    2000-12-01

    A survey has been performed to assess the numbers of all types of radiological x-ray examination conducted in the UK during the period from April 1997 to March 1998. The survey covers all diagnostic and interventional procedures using x-rays for medical and dental purposes, both within and outside the National Health Service (NHS), but excludes a detailed analysis of magnetic resonance imaging (MRI), ultrasound and nuclear medicine. This is the first such national survey conducted by NRPB since 1983. The results provide a current picture of the pattern of medical x-ray imaging practice in the UK and will allow revised estimates to be made of the collective dose to the population from these procedures. The survey has utilised detailed information available from radiology management systems at a selected sample of 38 English NHS trusts. The different classifications of x-ray procedure have been re-arranged into 62 standardised categories based on anatomical location and whether they were conventional, computed tomography (CT) or interventional procedures. Extrapolation of the sample data to the whole of England was carried out using broad NHS radiology statistics (KH12 returns) for the period of the survey from the Department of Health. Additional data have been obtained covering NHS radiology practice in Wales and Northern Ireland and also for x-ray imaging practice outside NHS hospitals such as that performed in independent hospitals and by dentists and chiropractors. Results are presented giving the annual numbers and relative frequencies of x-ray examinations in the 62 categories and the contributions from radiology practice outside NHS hospitals and from the whole of the UK. Altogether, about 41.5 million medical and dental x-ray examinations were conducted in the UK in 1997/98, corresponding to 704 examinations per 1000 inhabitants. The increase since 1983 for medical examinations conducted in NHS hospitals has just kept pace with the increase in population

  11. Control area around dental x-ray units - dosimetric study I

    International Nuclear Information System (INIS)

    Suric Mihic, M.; Prlic, I.; Milkovic-Kraus, S.; Mestrovic, T.; Rojnica, F.

    2005-01-01

    The issue of prompt professional occupational dose reporting is raised when the interval between doses is short or when the radiation source suffers a technical failure. Every involved person should be able to recognised individual or group radiation exposure. Actual radiation quality of the source is to be taken into account. To optimise radiation protection of dental radiologists, dental x-ray units were subject to Quality Control measurements. Scattering radiation from the patient's dental structures was measured in order to prove the results published by S. Tabakov, but using the modern RVG dental mode and several classical diagnostic positions. We used a special head phantom (real scull + Perspex + crown glass) and common dental x-ray units of various brands and types. The radiation quality was measured using standard QA/QC equipment. We measured the radiation scattered from the phantom in the horizontal plane (at thyroid height) at 0.5 m distance from the centre of the phantom. The measurement were done for a number of standard dental x-ray procedures, but this paper presents only the scattering caused by the upper premolars. The attenuation in the facial tissue was minimal and the majority of incidental radiation passes through the open mouth of a patient directly into the room area causing occupational exposure. The results we obtained are consistent with earlier reports on patient dosimetry. Occupational exposure is much lower if a modern RVG technique is used and no radiation protection threshold is exceeded in relation to Croatian laws. Much more important is the fact that the need for protective equipment and shielding is smaller if QA warrants proper technical operation of the x-ray tube. The maintenance of dental units is essential and so is a proper training of staff using modern diagnostic techniques. The control area around the x-ray unit is to be calculated and established for every standard dental unit (this does not apply for panoramic x-rays

  12. Investigation of clinical conditions affecting quality control in a dental X-ray unit

    International Nuclear Information System (INIS)

    Akiyama, Hironori; Koseki, Takakazu; Itagaki, Keisuke; Yotsui, Yoritaka; Gamoh, Shoko; Shimizutani, Kimishige

    2011-01-01

    In this study, to clarify quality control issues in a dental X-ray unit, we surveyed the uniformity and safety of absorbed doses and exposure time from dental X-ray units, and evaluated image quality of intraoral radiography at Osaka Dental University Hospital. Measurements of dose and exposure time were carried out five times and the mean was calculated at twenty-two dental X-ray units in our university hospital using the standard parameters for the mandibular molar region in adults. Intraoral radiography for evaluation of image quality was taken using D-speed films under the standard parameters for the mandibular molar region in adults and E-speed films for pediatric patients. Evaluation of image quality was performed by three oral radiologists. Dose and exposure time differed significantly among units. Image quality of intraoral radiography using D-speed films was not significantly better than that of E-speed films. This study demonstrated that it is necessary to establish a uniform dose and exposure time and to change D-speed films to E-speed films when obtaining intraoral radiography in our hospital. (author)

  13. The irradiation action on human dental tissue by X-rays and electrons. A nanoindenter study

    Energy Technology Data Exchange (ETDEWEB)

    Fraenzel, Wolfgang [Halle-Wittenberg Univ., Halle (Germany). Dept. of Physics; Gerlach, Reinhard [Halle-Wittenberg Univ., Halle (Germany). Clinic of Radiation Therapy

    2009-07-01

    It is known that ionizing radiation is used in medicine for Roentgen diagnostics and for radiation therapy. The radiation interacts with matter, in particular with biological one, essentially by scattering, photoelectric effect, Compton effect and pair production. To what extent the biological material is changed thereby, depends on the type and the amount of radiation energy, on the dose and on the tissue constitution. In modern radiation therapy two different kinds of radiation are used: high energy X-rays and electron radiation. In the case of head-neck tumors the general practice is an irradiation with high energy X-rays with absorbed dose to water up to 70 Gy. Teeth destruction has been identified as a side effect during irradiation. In addition, damage to the salivary glands is often observed which leads to a decrease or even the complete loss of the salivary secretion (xerostomia). This study shows how the different energy and radiation types damage the tooth tissue. The effects of both, high X-ray energy and high energy electrons, on the mechanical properties hardness and elasticity of the human dental tissue are measured by the nanoindentation technique. We compare these results with the effect of the irradiation of low X-ray energy on the dental tissue. (orig.)

  14. The irradiation action on human dental tissue by X-rays and electrons. A nanoindenter study

    International Nuclear Information System (INIS)

    Fraenzel, Wolfgang; Gerlach, Reinhard

    2009-01-01

    It is known that ionizing radiation is used in medicine for Roentgen diagnostics and for radiation therapy. The radiation interacts with matter, in particular with biological one, essentially by scattering, photoelectric effect, Compton effect and pair production. To what extent the biological material is changed thereby, depends on the type and the amount of radiation energy, on the dose and on the tissue constitution. In modern radiation therapy two different kinds of radiation are used: high energy X-rays and electron radiation. In the case of head-neck tumors the general practice is an irradiation with high energy X-rays with absorbed dose to water up to 70 Gy. Teeth destruction has been identified as a side effect during irradiation. In addition, damage to the salivary glands is often observed which leads to a decrease or even the complete loss of the salivary secretion (xerostomia). This study shows how the different energy and radiation types damage the tooth tissue. The effects of both, high X-ray energy and high energy electrons, on the mechanical properties hardness and elasticity of the human dental tissue are measured by the nanoindentation technique. We compare these results with the effect of the irradiation of low X-ray energy on the dental tissue. (orig.)

  15. Association of brain cancer with dental x-rays and occupation in Missouri

    International Nuclear Information System (INIS)

    Neuberger, J.S.; Brownson, R.C.; Morantz, R.A.; Chin, T.D.

    1991-01-01

    This investigation of a brain cancer cluster in Missouri used two approaches to investigate associations with potential risk factors. In a case-control study in a rural town, we interviewed surrogates of cases and controls about potential risk factors. We found a statistically significant positive association of brain cancer with reported exposure to dental x-rays. Occupation was not associated with the cluster in the rural town. In a standardized proportional mortality study for the state of Missouri, we calculated the observed and expected proportion of brain cancers by occupation and industry in Missouri decedents. We found that motor vehicle manufacturers, beauty shop workers, managers and administrators, elementary school teachers, and hairdressers and cosmetologists had significantly elevated proportions of brain cancer. Brain tumors are inconsistently associated with occupation in the literature. Further study of brain cancer etiology with respect to dental x-ray exposures seems warranted

  16. Parallelized Bayesian inversion for three-dimensional dental X-ray imaging.

    Science.gov (United States)

    Kolehmainen, Ville; Vanne, Antti; Siltanen, Samuli; Järvenpää, Seppo; Kaipio, Jari P; Lassas, Matti; Kalke, Martti

    2006-02-01

    Diagnostic and operational tasks based on dental radiology often require three-dimensional (3-D) information that is not available in a single X-ray projection image. Comprehensive 3-D information about tissues can be obtained by computerized tomography (CT) imaging. However, in dental imaging a conventional CT scan may not be available or practical because of high radiation dose, low-resolution or the cost of the CT scanner equipment. In this paper, we consider a novel type of 3-D imaging modality for dental radiology. We consider situations in which projection images of the teeth are taken from a few sparsely distributed projection directions using the dentist's regular (digital) X-ray equipment and the 3-D X-ray attenuation function is reconstructed. A complication in these experiments is that the reconstruction of the 3-D structure based on a few projection images becomes an ill-posed inverse problem. Bayesian inversion is a well suited framework for reconstruction from such incomplete data. In Bayesian inversion, the ill-posed reconstruction problem is formulated in a well-posed probabilistic form in which a priori information is used to compensate for the incomplete information of the projection data. In this paper we propose a Bayesian method for 3-D reconstruction in dental radiology. The method is partially based on Kolehmainen et al. 2003. The prior model for dental structures consist of a weighted l1 and total variation (TV)-prior together with the positivity prior. The inverse problem is stated as finding the maximum a posteriori (MAP) estimate. To make the 3-D reconstruction computationally feasible, a parallelized version of an optimization algorithm is implemented for a Beowulf cluster computer. The method is tested with projection data from dental specimens and patient data. Tomosynthetic reconstructions are given as reference for the proposed method.

  17. Application of digital tomosynthesis (DTS) of optimal deblurring filters for dental X-ray imaging

    International Nuclear Information System (INIS)

    Oh, J. E.; Cho, H. S.; Kim, D. S.; Choi, S. I.; Je, U. K.

    2012-01-01

    Digital tomosynthesis (DTS) is a limited-angle tomographic technique that provides some of the tomographic benefits of computed tomography (CT) but at reduced dose and cost. Thus, the potential for application of DTS to dental X-ray imaging seems promising. As a continuation of our dental radiography R and D, we developed an effective DTS reconstruction algorithm and implemented it in conjunction with a commercial dental CT system for potential use in dental implant placement. The reconstruction algorithm employed a backprojection filtering (BPF) method based upon optimal deblurring filters to suppress effectively both the blur artifacts originating from the out-focus planes and the high-frequency noise. To verify the usefulness of the reconstruction algorithm, we performed systematic simulation works and evaluated the image characteristics. We also performed experimental works in which DTS images of enhanced anatomical resolution were successfully obtained by using the algorithm and were promising to our ongoing applications to dental X-ray imaging. In this paper, our approach to the development of the DTS reconstruction algorithm and the results are described in detail.

  18. [Standards and guidelines of radiation protection and safety in dental X-ray examinations].

    Science.gov (United States)

    Guo, X L; Li, G; Cheng, Y; Yu, Q; Wang, H; Zhang, Z Y

    2017-12-09

    With the rapid development of imaging technology, the application of dental imaging in diagnosis, treatment planning, intraoperative surgical navigation, monitoring of treatment or lesion development and assessment of treatment outcomes is playing an essential role in oral healthcare. The increased total number of dental X-ray examinations is accompanied by a relatively significant increase in collective dose to patients as well as to dental healthcare workers, which is harmful to human bodies to a certain degree. Some radiation protection standards and guidelines in dental radiology have been published in European countries, US, Canada and Australia, etc. Adherence to these standards and guidelines helps to achieve images with diagnostic quality and avoid unnecessary and repeated exposures. However, no radiation protection standard or guideline with regard to dental X-ray examinations has been put in force so far in mainland China. Therefore, a literature review on available radiation protection standards and guidelines was conducted to provide reference to the development of radiation protection standards or guidelines in mainland China.

  19. Evaluation of the shielding of dental X-rays units; Evaluacion del blindaje de unidades de rayos X dentales

    Energy Technology Data Exchange (ETDEWEB)

    Medrano, E.; Vega C, H. R.; Letechipia de L, C.; Hernandez D, V. M.; Salas L, M. A., E-mail: edumeco@yahoo.com.mx [Universidad Autonoma de Zacatecas, Apdo. Postal 336, 98000 Zacatecas (Mexico)

    2014-08-15

    The capacity of the walls of the dental radio-diagnostic rooms has been determined, to diminish the dose levels during the use of the X-rays equipment s. The study was carried out in the Dentistry Academic Unit of the campus Siglo X XI of the Universidad Autonoma de Zacatecas. The X-rays equipment s are a learning tool for the dentistry students and they are also used for offering health services to the population; for this reason is important to verify that the dose levels outside of the room walls are safe. During the evaluation process were used conservative approaches without prejudice of the thickness necessary in benefit of the radiological protection. Of the evaluation was found that all the walls satisfy their function thoroughly like barriers against the X-rays. (Author)

  20. Device for measuring the exposure time in dental X-ray - Cronox

    International Nuclear Information System (INIS)

    Menezes, Claudio J.M.; Santos, Luiz A.P. dos

    2009-01-01

    The Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE) developed a test device for monitoring the X-ray beam in dental equipment to its application in quality control programs. This device, called Odontologic Dosimetric Card (CDO of Cartao Dosimetrico Odontologico in Portuguese) uses thermoluminescent dosimeters (TLD) for the measurement of some parameters of the X-ray beam as the entrance surface dose, the peak tension and half value layer (HVL). Radiographic films record the size of the radiation field. However, the TLD does not allow the assessment of exposure time, a parameter that complements the requirements of the Diretrizes de Protecao Radiologica em Radiodiagnostico Medico e Odontologico of Department of Health in Brazil for such equipment. Thus was developed a system based on sensitivity to ionizing radiation of phototransistors for measurement of exposure time when a patient is put in a clinical dental radiography. The system, called CRONOX was sized to be inserted within the CDO. The results showed that the measuring error had developed for less than 3% when compared to reference values obtained with the Tektronix digital oscilloscope, TDS2022 model. The readings obtained with the CRONOX were also compared with the nominal values selected in the X-ray equipment and with the values measured with the instrument of trade PTW Diavolt Universal. The results showed that the measuring device developed showed a maximum deviation of 5.92% on the nominal value selected, while for the instrument of PTW was 17.86%. (author)

  1. The survey of the surface doses of the dental x-ray machines

    International Nuclear Information System (INIS)

    Lee, Jae Seo; Kang, Byung Cheol; Yoon, Suk Ja

    2005-01-01

    The purpose of this study was to investigate variability of doses with same exposure parameters and evaluate radiographic density according to the variability of doses. Twenty-eight MAX-GLS (Shinhung Co, Seoul, Korea), twenty-one D-60-S (DongSeo Med, Seoul, Korea), and eleven REX-601 (Yoshida Dental MFG, Tokyo, Japan) dental x-ray machines were selected for this study. Surface doses were measured under selected combinations of tube voltage, tube current, exposure time, and constant distance 42 cm from the focal spot to the surface of the Multi-O-meter (Unfors Instrument, Billdal, Sweden). Radiographic densities were measured on the films at maximum, minimum and mean surface doses of each brand of x-ray units. With MAX-GLS, the maximum surface doses were thirteen to fourteen times as much as the minimum surfaces doses. With D-60-S, the maximum surface doses were three to eight times as much as the minimum surface doses. With REX-601, the maximum surface doses were six to ten times as much as the minimum surface doses. The differences in radiographic densities among maximum, mean, and minimum doses were significant (p<0.01). The surface exposure doses of each x-ray machine at the same exposure parameters were different within the same manufacturer's machines.

  2. Asbestos, dental x-rays, tobacco, and alcohol in the epidemiology of laryngeal cancer

    International Nuclear Information System (INIS)

    Hinds, M.W.; Thomas, D.B.; O'Reilly, H.P.

    1979-01-01

    A case-control study of 47 laryngeal cancers in males of three counties of Washington State was conducted. Personal interview was used to obtain information on smoking, alcohol use, exposure to asbestos, and other substances, and x-rays of the head and neck area. Smoking and alcohol consumption were found to increase risk of laryngeal cancer independently, with a clear dose-response relationship. Neither asbestos exposure nor exposure to other substances was found to significantly increase the risk of laryngeal cancer, although the relative risk with asbestos exposure was 1.75. Lifetime history of exposure to dental x-rays on five or more occasions was associated with significantly increased risk of laryngeal cancer among heavy smokers but not among light smokers. The importance of tobacco and alcohol in the epidemiology of laryngeal cancer was re-affirmed, the importance of asbestos exposure was brought into question, and a possible relationship of laryngeal cancer with exposure to dental x-rays among heavy smokers was demonstrated

  3. Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields.

    Science.gov (United States)

    Kelaranta, Anna; Ekholm, Marja; Toroi, Paula; Kortesniemi, Mika

    2016-01-01

    Dental radiography may involve situations where the patient is known to be pregnant or the pregnancy is noticed after the X-ray procedure. In such cases, the radiation dose to the foetus, though low, needs to be estimated. Uniform and widely used guidance on dental X-ray procedures during pregnancy are presently lacking, the usefulness of lead shields is unclear and practices vary. Upper estimates of radiation doses to the foetus and breasts of the pregnant patient were estimated with an anthropomorphic female phantom in intraoral, panoramic, cephalometric and CBCT dental modalities with and without lead shields. The upper estimates of foetal doses varied from 0.009 to 6.9 μGy, and doses at the breast level varied from 0.602 to 75.4 μGy. With lead shields, the foetal doses varied from 0.005 to 2.1 μGy, and breast doses varied from 0.002 to 10.4 μGy. The foetal dose levels without lead shielding were dental radiographic examination.

  4. Effect of dental metal in 10 MV X-ray beam therapy

    International Nuclear Information System (INIS)

    Mimura, Seiichi; Mikami, Yasutaka; Inamura, Keiji; Tahara, Seiji; Nagaya, Isao; Egusa, Tomomi; Nakagiri, Yoshitada; Sugita, Katsuhiko.

    1991-01-01

    We have often encountered patients with dental metal when employing the 10 MV X-ray beam therapy for head and neck tumors, and felt it important to investigate the effect of dental metal in relation to dose distribution. The absorbed dose rose abruptly in the vicinity of the metal reaching an interface value equal to 150% of the dose within the acrylic phantom. These results showed that an overdose occurred about 5 mm from the metal. We also learned that the overdose can be avoided by using a 5-mm thick tissue equivalent material. Six patients with dental metal were treated after first covering their metal with a 5-mm thick mouthpiece. No radiation stomatitis caused by the metal was observed in any of these cases. (author)

  5. Application of X-rays to dental age estimation in medico-legal practice

    Directory of Open Access Journals (Sweden)

    Dorota Lorkiewicz-Muszyńska

    2015-05-01

    Full Text Available Aim of the study: The paper addresses the use of dental age assessment methods based on radiographs in medico-legal practice. Different cases of practical application of the methods are presented including identification of human remains, dental age assessment in a living person and one archaeological case. Material and methods : The study material consisted of cases involving dental age assessment performed in the Department of Forensic Medicine, Poznan University of Medical Sciences in Poznan. Depending on the preliminary assessment of age, the Liversidge or the Kvaal et al. methods were applied. Dental age was estimated on the basis of available pantomograms. In the case of the living person, it was a radiograph supplied for expert evaluation. In the other cases, dental computed tomography was performed. Results : Dental age was successfully estimated in all of the cases. Various methods based on the analysis of X-ray images were applied. Dental age was shown to be correlated with skeletal age. Conclusions : The methods based on radiographs were demonstrated to be useful, and the results they yield are fully correlated with results of anthropological analyses.

  6. Radiation protection in dental X-ray surgeries--still rooms for improvement.

    Science.gov (United States)

    Hart, G; Dugdale, M

    2013-03-01

    To illustrate the authors' experience in the provision of radiation protection adviser (RPA)/medical physics expert (MPE) services and critical examination/radiation quality assurance (QA) testing, to demonstrate any continuing variability of the compliance of X-ray sets with existing guidance and of compliance of dental practices with existing legislation. Data was collected from a series of critical examination and routine three-yearly radiation QA tests on 915 intra-oral X-ray sets and 124 panoramic sets. Data are the result of direct measurements on the sets, made using a traceably calibrated Unfors Xi meter. The testing covered the measurement of peak kilovoltage (kVp); filtration; timer accuracy and consistency; X-ray beam size; and radiation output, measured as the entrance surface dose in milliGray (mGy) for intra-oral sets and dose-area product (DAP), measured in mGy.cm(2) for panoramic sets. Physical checks, including mechanical stability, were also included as part of the testing process. The Health and Safety Executive has expressed concern about the poor standards of compliance with the regulations during inspections at dental practices. Thirty-five percent of intra-oral sets exceeded the UK adult diagnostic reference level on at least one setting, as did 61% of those with child dose settings. There is a clear advantage of digital radiography and rectangular collimation in dose terms, with the mean dose from digital sets 59% that of film-based sets and a rectangular collimator 76% that of circular collimators. The data shows the unrealised potential for dose saving in many digital sets and also marked differences in dose between sets. Provision of radiation protection advice to over 150 general dental practitioners raised a number of issues on the design of surgeries with X-ray equipment and critical examination testing. There is also considerable variation in advice given on the need (or lack of need) for room shielding. Where no radiation protection

  7. Summary of: radiation protection in dental X-ray surgeries--still rooms for improvement.

    Science.gov (United States)

    Walker, Anne

    2013-03-01

    To illustrate the authors' experience in the provision of radiation protection adviser (RPA)/medical physics expert (MPE) services and critical examination/radiation quality assurance (QA) testing, to demonstrate any continuing variability of the compliance of X-ray sets with existing guidance and of compliance of dental practices with existing legislation. Data was collected from a series of critical examination and routine three-yearly radiation QA tests on 915 intra-oral X-ray sets and 124 panoramic sets. Data are the result of direct measurements on the sets, made using a traceably calibrated Unfors Xi meter. The testing covered the measurement of peak kilovoltage (kVp); filtration; timer accuracy and consistency; X-ray beam size; and radiation output, measured as the entrance surface dose in milliGray (mGy) for intra-oral sets and dose-area product (DAP), measured in mGy.cm(2) for panoramic sets. Physical checks, including mechanical stability, were also included as part of the testing process. The Health and Safety Executive has expressed concern about the poor standards of compliance with the regulations during inspections at dental practices. Thirty-five percent of intra-oral sets exceeded the UK adult diagnostic reference level on at least one setting, as did 61% of those with child dose settings. There is a clear advantage of digital radiography and rectangular collimation in dose terms, with the mean dose from digital sets 59% that of film-based sets and a rectangular collimator 76% that of circular collimators. The data shows the unrealised potential for dose saving in many digital sets and also marked differences in dose between sets. Provision of radiation protection advice to over 150 general dental practitioners raised a number of issues on the design of surgeries with X-ray equipment and critical examination testing. There is also considerable variation in advice given on the need (or lack of need) for room shielding. Where no radiation protection

  8. Hygiene implications associated with x-ray exposures to dental patients

    International Nuclear Information System (INIS)

    McKlveen, J.W.

    1980-01-01

    An elastic mask worn by patients, then a skeleton encased in plastic, was instrumented with LiF thermoluminescent dosimeters to quantify radiation exposures delivered from full-face diagnostic dental x-rays. Locations of interest included skin surface, eyes, upper and lower teeth and thyroid. Exposures in the 100 mR range were common and a maximum of over 6000 mR was measured in the teeth region during a full-face examination with a periapical unit. In general, exposures received from periapical equipment were several times those obtained from panoramic devices

  9. Radiation exposure of the UK population from medical and dental x-ray examinations

    International Nuclear Information System (INIS)

    Hart, D.; Wall, B.F.

    2002-03-01

    Knowledge of recent trends in the radiation doses from x-ray examinations and their distribution for the UK population provides useful guidance on where best to concentrate efforts on patient dose reduction in order to optimise the protection of the population in a cost-effective manner. In this report, the results of a recent survey of the frequency of medical and dental x-ray examinations in the UK and contemporary data on the radiation doses typically received by patients, are used to assess trends in the extent and the pattern of the population exposure. Individual patient doses, expressed in terms of the effective dose, range from a few microsieverts for simple radiographic examinations of the teeth, limbs or chest to tens of millisieverts for prolonged fluoroscopic procedures or some computed tomography (CT) examinations. A total of about 41.5 million medical and dental x-ray examinations are now conducted each year in the UK (0.70 examination per head of population) resulting in an annual per caput effective dose of 330 μSv. This is not significantly different from the previous rough estimate of 350 μSv for 1991. However, over the last ten years CT has more than doubled its contribution and is now responsible for 40% of the total dose to the population from medical x-rays. In contrast, the contribution from conventional radiographic and fluoroscopic examinations has nearly halved to about 44%. Interventional and angiographic procedures together contribute the remaining 16%. The annual per caput dose of 330 μSv is low in comparison with other countries having similarly developed systems of health care. This is due to both a lower frequency of x-ray examinations per head of population and generally lower doses in the UK than in other developed countries. However, the much increased contributions of CT, angiography and interventional procedures to the UK population dose indicate an urgent need to develop radiation protection and optimisation activities for

  10. Development of a concept for radiation patients exposure assessment during dental X-ray examinations and statistical data acquisition for the determination of a diagnostic reference value

    International Nuclear Information System (INIS)

    Kueppers, C.; Sering, M.; Poppe, B.; Poplawski, A.; Looe, H.K.; Beyer, D.; Pfaffenberger, A.; Chofor, N.; Eenboom, F.

    2012-01-01

    The research project on the development a concept for radiation patients exposure assessment during dental X-ray examinations and statistical data acquisition for the determination of a diagnostic reference value includes the following issues: Fundamental facts: dental X-ray examination techniques, dose relevant factors and characteristics during X-ray examinations, radiation exposed organs during dental X-ray examinations, dose assessment based on phantoms. Materials and methodologies of the project: TLD measurements using the phantom, calculation of the effective dose during dental X-ray examinations, properties and settings of the reference facilities for the determination of radiation exposure, selection of dental offices, dosimetric measurements, data acquisition and statistical evaluation. Results of dosimetric examinations: results of dosimetric measurements at reference facilities, results of dosimetric measurements in dental offices. Discussion of the concept for the determination of the radiation exposure during dental X-ray examinations.

  11. The dose received by patients during dental X-ray examination and the technical condition of radiological equipment.

    Science.gov (United States)

    Bekas, Marcin; Pachocki, Krzysztof A

    2013-01-01

    Implementation of X-ray dental examination is associated with the patients exposure to ionizing radation. The size of the exposure depends on the type of medical procedure, the technical condition of the X-ray unit and selected exposure conditions. The aim of this study was to determine the dose received by patients during dental X-ray examination and the assessment of the technical condition of medical equipment, The study included a total number of 79 dental X-ray units located in the region of Mazovia. The test methods for the assessment of the technical condition of dental X-ray units and measurement of radiation dose received by patients were based on the procedures elaborated in the Department of Radiation Hygiene and Radiobiology in the National Institute of Public Health - National Institute of Hygiene (Warszawa, Poland) accredited for the certification of compliance with PN-EN 17025. The research found that 69.6% fully meets the criteria set out in the Polish legislation regarding the safe use of ionizing radiation in medicine, while 30.4% did not meet some of them. A tenfold difference in the size of the dose received by patients during dental X-ray examinations was discovered. For example, during a radiography of the canine teeth of a child, the recorded entrance surface dose (ESD) ranged from 72.8 to 2430 microGy with the average value of 689.1 microGy. Cases where the dose reference level defined in Polish legislation of 5 mGy was exceeded were also found. CONCKUSIONS: It is essential to constantly monitor the situation regarding the technical condition of X-ray units which affects the size of the population's exposure to ionizing radiation as well as raising dentists' awareness about the effects of X-rays on the human body.

  12. The reliability of dental x-ray film in assessment of MP3 stages of the pubertal growth spurt.

    Science.gov (United States)

    Abdel-Kader, H M

    1998-10-01

    The main object of this clinical study is to provide a simple and practical method to assess the pubertal growth spurt stages of a subject by recording MP3 stages with the dental periapical radiograph and the standard dental x-ray machine.

  13. Dose distribution in head and neck during dental x-ray procedures

    International Nuclear Information System (INIS)

    Mason, E.W.; Goepp, R.A.

    1978-01-01

    Previous studies, notably by Franklin (Angle Ortho., 43:53-64, 1973), have shown significant exposures during cephalometric dental procedures and ways in which these exposures can be reduced. Skin dose over thyroid tissue has been measured by Alcox (J. Am. Dent. Assoc., 88:568-579, 1974), and others. This study is an expansion of thyroid dose measurements by Block, Goepp, and Mason (Angle Ortho., 47:17-24, 1977). The internal dose distribution in the head and neck area due to cephalometric and panoramic dental x-ray procedures is shown along with the dependence of orbit and thyroid dose on patient positioning. Higher doses can be delivered to deep tissue by panoramic machines since tissue at the axis of rotation is exposed during the entire procedure. (author)

  14. X-ray scattering evaluation of ultrastructural changes in human dental tissues with thermal treatment.

    Science.gov (United States)

    Sandholzer, Michael A; Sui, Tan; Korsunsky, Alexander M; Walmsley, Anthony Damien; Lumley, Philip J; Landini, Gabriel

    2014-05-01

    Micro- and ultrastructural analysis of burned skeletal remains is crucial for obtaining a reliable estimation of cremation temperature. Earlier studies mainly focused on heat-induced changes in bone tissue, while this study extends this research to human dental tissues using a novel quantitative analytical approach. Twelve tooth sections were burned at 400-900°C (30-min exposure, increments of 100°C). Subsequent combined small- and wide-angle X-ray scattering (SAXS/WAXS) experiments were performed at the Diamond Light Source synchrotron facility, where 28 scattering patterns were collected within each tooth section. In comparison with the control sample, an increase in mean crystal thickness was found in burned dentine (2.8-fold) and enamel (1.4-fold), however at a smaller rate than reported earlier for bone tissue (5-10.7-fold). The results provide a structural reference for traditional X-ray scattering methods and emphasize the need to investigate bone and dental tissues separately to obtain a reliable estimation of cremation temperature. © 2014 American Academy of Forensic Sciences.

  15. Synchrotron-radiation-based X-ray micro-computed tomography reveals dental bur debris under dental composite restorations.

    Science.gov (United States)

    Hedayat, Assem; Nagy, Nicole; Packota, Garnet; Monteith, Judy; Allen, Darcy; Wysokinski, Tomasz; Zhu, Ning

    2016-05-01

    Dental burs are used extensively in dentistry to mechanically prepare tooth structures for restorations (fillings), yet little has been reported on the bur debris left behind in the teeth, and whether it poses potential health risks to patients. Here it is aimed to image dental bur debris under dental fillings, and allude to the potential health hazards that can be caused by this debris when left in direct contact with the biological surroundings, specifically when the debris is made of a non-biocompatible material. Non-destructive micro-computed tomography using the BioMedical Imaging & Therapy facility 05ID-2 beamline at the Canadian Light Source was pursued at 50 keV and at a pixel size of 4 µm to image dental bur fragments under a composite resin dental filling. The bur's cutting edges that produced the fragment were also chemically analyzed. The technique revealed dental bur fragments of different sizes in different locations on the floor of the prepared surface of the teeth and under the filling, which places them in direct contact with the dentinal tubules and the dentinal fluid circulating within them. Dispersive X-ray spectroscopy elemental analysis of the dental bur edges revealed that the fragments are made of tungsten carbide-cobalt, which is bio-incompatible.

  16. Development of a dosimetric system for dental X-ray equipment quality control

    International Nuclear Information System (INIS)

    Melo, Francisco Almeida de

    2002-08-01

    An electronic instrument with digital readout was designed and constructed to provide fast, simple and non-invasive measurements of X-ray dental equipment parameters. This instrument is capable of evaluating the entrance dose, exposure time, tube voltage (kVp) and beam filtration. It consists of a set of five photodiodes connected each one in the photocurrent mode to the input of a designed integrating electrometer. Three of the detectors are fixed under aluminium filters with different thicknesses, one is fixed under a 0.3 mm copper filter and the other has no filtration. The readings of the three detectors under aluminium filters and of the bare detector permit the determination of the half-value layer, which is used to calculate the beam filtration. The ratio between the readings of the detector below the copper filter and the one without filtration is used to determine the tube voltage. The signal produced by the detector without filter is used to evaluate the patient entrance dose, and to active an electronic timer for measuring the real exposure time. The tests and calibration of the instrument in different voltages in the 59 - 70kVp range, showed that its response is both stable and reproducible to within 1%. The instrument response was compared to the one from a commercial non-invasive X-ray test equipment (Gammex RMI Multifunction kVp meter). The results showed that the response of the developed instrument is in good agreement with the RMI meter which is the standard equipment for such measurements. These results indicate that the dosimetric system is suitable for use in Dental Quality Assurance Programs. (author)

  17. A study on developpement of guideline on writing technical document for electrical medical devices: Dental x-ray equipment

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Seung Youl; Kim, Jae Ryang; Lee, Jun Ho; Park, Chang Won [Division of Medical Device Research, National Institute of Food and Drug Safety Evaluation, Mnistry of Food and Drug Safety (Korea, Republic of)

    2016-12-15

    Due to recent population aging, the number of check-up for senior citizens has increased steadily. According to this trend, the market size of dental X-ray equipment and the number of approval and review for these devices have simultaneously increased. The technical document of medical device is required for approval and review for medical device, and medical device companies needs to have work comprehension and expertise, as the document needs to include the overall contents such as performances, test criteria, etc.. Yet, since most of domestic manufacturers or importers of medical devices are small businesses, it is difficult for them to recruit professional manpower for approval of medical devices, and submission of inaccurate technical documents has increased. These problems lead to delay of the approval process and to difficulties in quick entering into the market. Especially, the Ministry of Food and Drug safety (MFDS) standards of a dental extra-oral X-ray equipment, a dental intra-oral X-ray equipment, an arm-type computed tomography, and a portable X-ray system have been recently enacted or not. this guideline of dental X-ray equipment adjusting revised standards was developed to help relative companies and reviewers. For this study, first, the methods to write technical document have been reviewed with revised international and domestic regulations and system. Second, the domestic and foreign market status of each item has been surveyed and analyzed. Third, the contents of technical documents already approved by MFDS have been analyzed to select the correct example, test items, criteria, and methods. Finally, the guideline has been developed based on international and domestic regulation, through close review of a consultative body composed of academic, industrial, research institute and government experts.

  18. A study on developpement of guideline on writing technical document for electrical medical devices: Dental x-ray equipment

    International Nuclear Information System (INIS)

    Lee, Seung Youl; Kim, Jae Ryang; Lee, Jun Ho; Park, Chang Won

    2016-01-01

    Due to recent population aging, the number of check-up for senior citizens has increased steadily. According to this trend, the market size of dental X-ray equipment and the number of approval and review for these devices have simultaneously increased. The technical document of medical device is required for approval and review for medical device, and medical device companies needs to have work comprehension and expertise, as the document needs to include the overall contents such as performances, test criteria, etc.. Yet, since most of domestic manufacturers or importers of medical devices are small businesses, it is difficult for them to recruit professional manpower for approval of medical devices, and submission of inaccurate technical documents has increased. These problems lead to delay of the approval process and to difficulties in quick entering into the market. Especially, the Ministry of Food and Drug safety (MFDS) standards of a dental extra-oral X-ray equipment, a dental intra-oral X-ray equipment, an arm-type computed tomography, and a portable X-ray system have been recently enacted or not. this guideline of dental X-ray equipment adjusting revised standards was developed to help relative companies and reviewers. For this study, first, the methods to write technical document have been reviewed with revised international and domestic regulations and system. Second, the domestic and foreign market status of each item has been surveyed and analyzed. Third, the contents of technical documents already approved by MFDS have been analyzed to select the correct example, test items, criteria, and methods. Finally, the guideline has been developed based on international and domestic regulation, through close review of a consultative body composed of academic, industrial, research institute and government experts

  19. Clinical usefulness of dental X-ray computed tomography for postoperative assessment of secondary alveolar bone grafting

    International Nuclear Information System (INIS)

    Noguchi, Kazuhide; Hamada, Yoshiki; Kondoh, Toshirou; Ishii, Hiroaki; Sonoyama, Tomoo; Kawarada, Takashi; Seto, Kanichi

    2003-01-01

    In this study, the clinical usefulness of dental X-ray computed tomography (CT) for postoperative assessment of secondary alveolar bone grafting was investigated. Nineteen bone-grafted alveolar clefts in 15 patients with cleft lip and palate were studied. All bone bridges were examined by dental three-dimensional (3D)-CT (PSR 9000: Asahi Roentgen, Kyoto, Japan). The postoperative 3D morphology of the bone bridges was easily recognized. Dental 3D-CT images were suggested to be useful for assessment before installation of dental implants in bone bridges. In addition, the status of bone surrounding the installed dental implants and the periodontal space of teeth adjacent to the cleft could be clearly evaluated. In conclusion, dental 3D-CT provides clinically valuable information for the postoperative assessment of secondary alveolar bone grafting. (author)

  20. Study on the characteristics of insight dental x-ray film

    International Nuclear Information System (INIS)

    Song, Young Han; Lee, Wan; Lee, Byung Do

    2003-01-01

    To investigate the characteristics of the newly marketed, Insight dental X-ray film. Kodak Ultraspeed (DF-58), E-speed, Agfa Dentus M2, and Kodak Insight (IP-21) films were radiographed using a Trophy intra-oral radiographic machine. 10 step exposure times were prepared and each step exposure was monitored using a FH 40G (ESM Eberline Instruments) dosimeter for each of the 4 types of intra-oral film. All films were manually processed and the radiographic densities at 6 site of each processed film were measured, and the characteristic curves of each of the 4 types intra-oral films were created utilizing these dosimetric data and radiographic densities, based on ISO 5779. The film contrast, speed, and base plus fog density of Insight film were compared with those of the 3 other films examined in this experiment. E-speed film showed greatest average gradients followed by Insight film. E-speed and Ultraspeed film showed great average gradients at low density levels. Insight film showed the fastest speed followed by E-speed, Dentus M2 and Ultraspeed film. Dentus M2 film showed greatest base plus fog density level followed by Insight film. Kodak Insight film showed fastest film speed with comparable film contrast on characteristic curve.

  1. Dental X-ray exposure and Alzheimer's disease: a hypothetical etiological association.

    Science.gov (United States)

    Rodgers, Caroline C

    2011-07-01

    Despite the fact that Alzheimer's disease was identified more than 100 years ago, its cause remains elusive. Although the chance of developing Alzheimer's disease increases with age, it is not a natural consequence of aging. This article proposes that dental X-rays can damage microglia telomeres - the structures at the end of chromosomes that determine how many times cells divide before they die - causing them to age prematurely. Degenerated microglia lose their neuroprotective properties, resulting in the formation of neurofibrillary tau tangles and consequently, the neuronal death that causes Alzheimer's dementia. The hypothesis that Alzheimer's is caused specifically by microglia telomere damage would explain the delay of one decade or longer between the presence of Alzheimer's brain pathology and symptoms; telomere damage would not cause any change in microglial function, it would just reset the countdown clock so that senescence and apoptosis occurred earlier than they would have without the environmental insult. Once microglia telomere damage causes premature aging and death, the adjacent neurons are deprived of the physical support, maintenance and nourishment they require to survive. This sequence of events would explain why therapies and vaccines that eliminate amyloid plaques have been unsuccessful in stopping dementia. Regardless of whether clearing plaques is beneficial or harmful - which remains a subject of debate - it does not address the failing microglia population. If microglia telomere damage is causing Alzheimer's disease, self-donated bone marrow or dental pulp stem cell transplants could give rise to new microglia populations that would maintain neuronal health while the original resident microglia population died. Copyright © 2011 Elsevier Ltd. All rights reserved.

  2. The development of a postal method to assess X-ray beam parameters and image quality in dental radiology

    International Nuclear Information System (INIS)

    Fenton, D.M.

    1994-10-01

    Intraoral radiographs are an extremely valuable diagnostic tool in dentistry. Radiography permits the early detection and diagnosis of dental disease and consequently is used extensively. However, public concern about radiation exposure has increased in recent times. This concern is reflected in national and international law, to the extent that, the basic principles of radiological protection, that is, justification, optimisation and dose limitation are written into law. Furthermore, in Ireland, the regulations, as outlined in the Code of Practice for Radiological Protection in Dentistry, require intraoral dental X-ray machines to perform to certain standards. A report of a direct survey of 164 intraoral dental X-ray machines is given in this study. The survey covered mechanical, electrical as well as radiation safety. Inadequacies with respect to focus to skin distance and timer accuracy were found in 45% and 42% of the machines surveyed. Ninety eight machines were assessed for electrical safety in which 48% were found to be unsafe. The results indicate that a complete assessment of the performance of dental X-ray units in Ireland is required. However, as there are in excess of 800 dental X-ray machines located throughout the country, such an assessment would be very costly for the regulatory authority. The development of a postal method for the assessment of the performance of dental X-ray machines is described in this study. This postal method provides information on the kV, total filtration, beam width and timer linearity and is undertaken by means of a penetrameter and film envelopes for exposure to the X-ray set under examination, together with a questionnaire that requests information on environment in which the machine is located. Using this method an accuracy of +-5% of the actual value was achieved in the measurement of kVp. The penetrameter was also used to assess whether or not the filtration of a particular machine complies with the regulations. This

  3. Development of a digital panoramic X-ray imaging system of adaptive image layers for dental applications

    International Nuclear Information System (INIS)

    Choi, S.I.; Park, Y.O.; Cho, H.S.; Oh, J.E.; Cho, H.M.; Hong, D.K.; Lee, M.S.; Yang, Y.J.; Je, U.K.; Kim, D.S.; Lee, H.K.

    2011-01-01

    As a continuation of our digital radiographic sensor R and D, we have developed a prototyped digital panoramic X-ray imaging system for dental applications. The imaging system consists of a slit-collimated X-ray generator with a 0.4 mm focal spot size and a 3.5 mm Al filtration, a linear-array typed CMOS imager with a 48x48 μm 2 pixel size and a 128 (in the scan direction)x3072 (in the vertical direction) pixel format, a series of microstep motors for the precise motion control of the imaging system, and the designed sequences for the motion control and pixel readout required to make a specific plane of interest (POI) to be focused. With the several test phantoms we designed, we obtained useful digital panoramic X-ray images by moving the X-ray generator and the CMOS imager along a continuously sliding rotational center. In this study, we demonstrated that the prototype system can be applicable to any shaped POI or multi-POIs simultaneously to be focused, provided that adequate sequences for motion control and pixel readout are designed. We expect that the imaging system will be useful for our ongoing applications of dental panoramic radiography and nondestructive testing.

  4. The study on the protection and actual condition of using the dental x-ray unit

    International Nuclear Information System (INIS)

    Kang, Eun Ju; Yoo, Beong Gyu

    2000-01-01

    This paper will present the result of research which was done with 201 places on the actual condition of using dental diagnostic radiography unit and the protection of radiography. The purpose of this paper is to comprehend the actual condition of using dental x-ray unit and to protect when they do radiation work. Moreover this paper was completed to prepare basic materials that could be helpful to reduce the exposure from radiation. This paper obtains the following result. On radiation photographing work in the dentist office, 60.3% of dental hygienists treat this job, and 19.2% of assistants, 10.8% of dentists, 5.6% of radiolotechnologists and 4.2% others performed this job. The case that radiation worker is educated about diagnostic radiography safety supervision has been shown 14.4% and uneducated case has been shown 78.1%. The result about the actual condition of using the oral diagnostic radiation per day was that a number of film which take photograph again (less than 1 exposure) was 40.3%. Normal photographing (1 ∼ 10 exposure) was 85.1% which is the highest percentage. Using the bitewing film and occlusal film was 7.0%, and 12.4% respectively. The percent that they use cephalo film and panoramic film was 16.4%, 29.8% respectively. Dental intra diagnostic radiography unit made in 1996 ∼ 2000 was 24.9% and the one made in 1991 ∼ 1995 was 19.9%, in 1986 ∼ 1990 was 19.9%, in 1985 was 9.5% according to the answer. On kVp, they use 60 kVp mostly (61.7%) and On mA, they use 10 mA with the highest percent (66.7%). On the dental extra diagnostic radiography units which are used for doing the extra oral radiography, the one made in 1996 ∼ 2000 was 13.4%, in 1991 ∼ 1995 was 9.5%, in 1985 ∼ 1990 was 2.0% according to the answer. They use 71 ∼ 80 kVp with 10.9% and 60 ∼ 75 kVp with 9.5%. They use less than 10 mA with 19.4% and 11 ∼ 15 mA with 2.5%, 16 ∼ 20 mA with 1.5%. But the case they exactly do not know how much mA they use or they do not have

  5. The study on the protection and actual condition of using the dental x-ray unit

    Energy Technology Data Exchange (ETDEWEB)

    Kang, Eun Ju; Yoo, Beong Gyu [Wonkwang Health Science College, Iksan (Korea, Republic of)

    2000-04-15

    This paper will present the result of research which was done with 201 places on the actual condition of using dental diagnostic radiography unit and the protection of radiography. The purpose of this paper is to comprehend the actual condition of using dental x-ray unit and to protect when they do radiation work. Moreover this paper was completed to prepare basic materials that could be helpful to reduce the exposure from radiation. This paper obtains the following result. On radiation photographing work in the dentist office, 60.3% of dental hygienists treat this job, and 19.2% of assistants, 10.8% of dentists, 5.6% of radiolotechnologists and 4.2% others performed this job. The case that radiation worker is educated about diagnostic radiography safety supervision has been shown 14.4% and uneducated case has been shown 78.1%. The result about the actual condition of using the oral diagnostic radiation per day was that a number of film which take photograph again (less than 1 exposure) was 40.3%. Normal photographing (1 {approx} 10 exposure) was 85.1% which is the highest percentage. Using the bitewing film and occlusal film was 7.0%, and 12.4% respectively. The percent that they use cephalo film and panoramic film was 16.4%, 29.8% respectively. Dental intra diagnostic radiography unit made in 1996 {approx} 2000 was 24.9% and the one made in 1991 {approx} 1995 was 19.9%, in 1986 {approx} 1990 was 19.9%, in 1985 was 9.5% according to the answer. On kVp, they use 60 kVp mostly (61.7%) and On mA, they use 10 mA with the highest percent (66.7%). On the dental extra diagnostic radiography units which are used for doing the extra oral radiography, the one made in 1996 {approx} 2000 was 13.4%, in 1991 {approx} 1995 was 9.5%, in 1985 {approx} 1990 was 2.0% according to the answer. They use 71 {approx} 80 kVp with 10.9% and 60 {approx} 75 kVp with 9.5%. They use less than 10 mA with 19.4% and 11 {approx} 15 mA with 2.5%, 16 {approx} 20 mA with 1.5%. But the case they

  6. Some properties of six different types of dental x-ray unit

    NARCIS (Netherlands)

    Trouerbach, W.Th.; Aken, J. van

    1965-01-01

    Forty-four x-ray machines were tested for reliability of timer and for output. A distinction was made between different types of errors, and large differences between these categories of errors were found.

  7. Low Dose X-Ray Sources and High Quantum Efficiency Sensors: The Next Challenge in Dental Digital Imaging?

    Directory of Open Access Journals (Sweden)

    Arnav R. Mistry

    2014-01-01

    Full Text Available Objective(s. The major challenge encountered to decrease the milliamperes (mA level in X-ray imaging systems is the quantum noise phenomena. This investigation evaluated dose exposure and image resolution of a low dose X-ray imaging (LDXI prototype comprising a low mA X-ray source and a novel microlens-based sensor relative to current imaging technologies. Study Design. A LDXI in static (group 1 and dynamic (group 2 modes was compared to medical fluoroscopy (group 3, digital intraoral radiography (group 4, and CBCT scan (group 5 using a dental phantom. Results. The Mann-Whitney test showed no statistical significance (α=0.01 in dose exposure between groups 1 and 3 and 1 and 4 and timing exposure (seconds between groups 1 and 5 and 2 and 3. Image resolution test showed group 1 > group 4 > group 2 > group 3 > group 5. Conclusions. The LDXI proved the concept for obtaining a high definition image resolution for static and dynamic radiography at lower or similar dose exposure and smaller pixel size, respectively, when compared to current imaging technologies. Lower mA at the X-ray source and high QE at the detector level principles with microlens could be applied to current imaging technologies to considerably reduce dose exposure without compromising image resolution in the near future.

  8. Theory of X-ray microcomputed tomography in dental research: application for the caries research

    OpenAIRE

    Young-Seok Park; Kwang-Hak Bae; Juhea Chang; Won-Jun Shon

    2011-01-01

    Caries remains prevalent throughout modern society and is the main disease in the field of dentistry. Although studies of this disease have used diverse methodology, recently, X-ray microtomography has gained popularity as a non-destructive, 3-dimensional (3D) analytical technique, and has several advantages over the conventional methods. According to X-ray source, it is classified as monochromatic or polychromatic with the latter being more widely used due to the high cost of the monochromat...

  9. Exposure to ionizing radiation during dental X-rays is not associated with risk of developing meningioma: a meta-analysis based on seven case-control studies.

    Science.gov (United States)

    Xu, Ping; Luo, Hong; Huang, Guang-Lei; Yin, Xin-Hai; Luo, Si-Yang; Song, Ju-Kun

    2015-01-01

    Many observational studies have found that exposure to dental X-rays is associated with the risk of development of meningioma. However, these findings are inconsistent. We conducted a meta-analysis to assess the relationship between exposure to dental X-rays and the risk of development of meningioma. The PubMed and EMBASE databases were searched to identify eligible studies. Summary odds ratio (OR) estimates and 95% confidence intervals (95% CIs) were used to compute the risk of meningioma development according to heterogeneity. Subgroup and sensitivity analyses were performed to further explore the potential heterogeneity. Finally, publication bias was assessed. Seven case-control studies involving 6,174 patients and 19,459 controls were included in the meta-analysis. Neither exposure to dental X-rays nor performance of full-mouth panorex X-rays was associated with an increased risk of development of meningioma (overall: OR, 0.97; 95% CI, 0.70-1.32; dental X-rays: OR, 1.05; 95% CI, 0.89-1.25; panorex X-rays: OR, 1.01; 95% CI, 0.76-1.34). However, exposure to bitewing X-rays was associated with a slightly increased risk of development of meningioma (OR, 1.73; 95% CI, 1.28-2.34). Similar results were obtained in the subgroup and sensitivity analyses. Little evidence of publication bias was observed. Based on the currently limited data, there is no association between exposure to dental X-rays and the risk of development of meningioma. However, these results should be cautiously interpreted because of the heterogeneity among studies. Additional large, high-quality clinical trials are needed to evaluate the association between exposure to dental X-rays and the risk of development of meningioma.

  10. Exposure to ionizing radiation during dental X-rays is not associated with risk of developing meningioma: a meta-analysis based on seven case-control studies.

    Directory of Open Access Journals (Sweden)

    Ping Xu

    Full Text Available Many observational studies have found that exposure to dental X-rays is associated with the risk of development of meningioma. However, these findings are inconsistent. We conducted a meta-analysis to assess the relationship between exposure to dental X-rays and the risk of development of meningioma.The PubMed and EMBASE databases were searched to identify eligible studies. Summary odds ratio (OR estimates and 95% confidence intervals (95% CIs were used to compute the risk of meningioma development according to heterogeneity. Subgroup and sensitivity analyses were performed to further explore the potential heterogeneity. Finally, publication bias was assessed.Seven case-control studies involving 6,174 patients and 19,459 controls were included in the meta-analysis. Neither exposure to dental X-rays nor performance of full-mouth panorex X-rays was associated with an increased risk of development of meningioma (overall: OR, 0.97; 95% CI, 0.70-1.32; dental X-rays: OR, 1.05; 95% CI, 0.89-1.25; panorex X-rays: OR, 1.01; 95% CI, 0.76-1.34. However, exposure to bitewing X-rays was associated with a slightly increased risk of development of meningioma (OR, 1.73; 95% CI, 1.28-2.34. Similar results were obtained in the subgroup and sensitivity analyses. Little evidence of publication bias was observed.Based on the currently limited data, there is no association between exposure to dental X-rays and the risk of development of meningioma. However, these results should be cautiously interpreted because of the heterogeneity among studies. Additional large, high-quality clinical trials are needed to evaluate the association between exposure to dental X-rays and the risk of development of meningioma.

  11. Evaluation of the effects of high energy X-ray radiation in materials used in dental restorations

    International Nuclear Information System (INIS)

    Maio, Mireia Florencio; Santos, Adimir dos; Fernandes, Marco Antonio Rodrigues

    2011-01-01

    This work studied the behavior of the physical features and chemical composition of materials used in dental restorations (titanium, amalgam, composite resin and glass ionomer cement) which were submitted to x-ray radiation of 6.0 Mega-Volt (MV) of energy produced in a linear accelerator that is used in radiotherapy of head and neck tumors 1 2. The samples were analyzed using a x-ray fluorescence technique by comparing the chemical composition before and after irradiation. In order to check the residual radiation in the samples, measurements of the sample dosimetry were performed with Geiger-Mueller radiation detectors and an ionization chamber. The samples were also analyzed by gamma-ray spectrometry using a hyper-pure Germanium (HPGe) detector. From these tests, we aimed to verify small changes in the composition of the test bodies due to the radiation. (author)

  12. Evaluation of the effects of high energy X-ray radiation in materials used in dental restorations

    Energy Technology Data Exchange (ETDEWEB)

    Maio, Mireia Florencio; Santos, Adimir dos, E-mail: mfmaio@ipen.br, E-mail: asantos@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Fernandes, Marco Antonio Rodrigues, E-mail: marfernandes@fmb.unesp.br [Universidade Estadual Paulista Julio de Mesquita Filho (UNESP), Botucatu, SP (Brazil). Departamento de Radioterapia

    2011-07-01

    This work studied the behavior of the physical features and chemical composition of materials used in dental restorations (titanium, amalgam, composite resin and glass ionomer cement) which were submitted to x-ray radiation of 6.0 Mega-Volt (MV) of energy produced in a linear accelerator that is used in radiotherapy of head and neck tumors 1 2. The samples were analyzed using a x-ray fluorescence technique by comparing the chemical composition before and after irradiation. In order to check the residual radiation in the samples, measurements of the sample dosimetry were performed with Geiger-Mueller radiation detectors and an ionization chamber. The samples were also analyzed by gamma-ray spectrometry using a hyper-pure Germanium (HPGe) detector. From these tests, we aimed to verify small changes in the composition of the test bodies due to the radiation. (author)

  13. Update of diagnostic medical and dental x-ray exposures in Romania

    Energy Technology Data Exchange (ETDEWEB)

    Sorop, Ioana; Mossang, Daniela; Dadulescu, Elena [Radiation Hygiene Laboratory of Public Health Authority Dolj, 2, Constantin Lecca Street, Craiova (Romania); Iacob, Mihai Radu [University ' Alexandru Ioan Cuza' , 11, Carol I Street, 700506, Iasi (Romania); Iacob, Olga [Institute of Public Health, 14, Victor Babes Street, 700465 Iasi (Romania)], E-mail: danamossang@sanpubdj.ro

    2008-12-15

    This national study, the third in the last 15 years, updates the magnitude of medical radiation exposure from conventional x-ray examinations, in order to optimise the radiological protection to the population in a cost-effective manner. Effective doses from diagnostic radiology were estimated for adult and paediatric patients undergoing the 20 most important types of x-ray examination. Data were collected from 179 x-ray departments, selected by their annual workload, throughout the country. Estimates were made using two dosimetric quantities: entrance surface dose, derived from the absorbed dose in air measured by simulation of radiographic examinations, and dose-area product, measured during fluoroscopic examinations performed on adult and paediatric patients. Conversion coefficients to effective dose of the UK National Radiological Protection Board (NRPB) have been used in all calculations. The effective dose per patient from all medical x-ray examinations was 0.74 mSv and the resulting annual collective effective dose was 6930 man Sv, with annual effective dose per caput of 0.33 mSv. The current size of population exposure from diagnostic radiology is lower than the previous one by 40%, but could be about 30% higher by taking into account the estimated contribution from computed tomography (CT) procedures.

  14. Hand-held and automated breast ultrasound

    International Nuclear Information System (INIS)

    Bassett, L.W.; Gold, R.H.; Kimme-Smith, C.

    1985-01-01

    The book is a guide for physicians and technologists who use US as an adjunct to mammography; it carefully outlines the pros and cons of US of the breast and its role in the diagnosis of benign and malignant diseases. After an introduction that discusses the philosophy of breast US, the chapters cover the physics of US and instrumentation (both hand-held transducers as well as automated water path scanners), then proceed to a discussion of the normal breast. Sections on benign disorders, malignant lesions, and pitfalls of diagnosis are followed by quiz cases

  15. The development and operation of a method for the remote determination of X-ray beam parameters used in dental radiography

    International Nuclear Information System (INIS)

    Hewitt, J.M.

    1984-07-01

    The method described is a part of the Dental Monitoring Service operated by the Board in the UK for the assessment of radiation protection in dental practice. This postal service, which provides a comprehensive survey of dental X-ray sets and radiographic procedures, is undertaken by means of a questionnaire, film cassettes for exposure to the X-ray set and a personal monitoring component to check operator doses. The film cassettes and the methods by which the X-ray beam parameters are obtained are described in detail. The cassettes use radiation monitoring film to realise, by means of measurements of relative transmission through selected copper filters, the extended dynamic range of exposure necessary for accurate indication of the operating kilovoltage and total beam filtration. The standard of the X-ray unit with regard to the relevant regulations and code of practice can then be assessed, and, from the values of radiation dose determined for chosen exposure times, exposure settings for optimum quality radiographs can be recommended where appropriate. Although designed primarily for dental X-ray units, use of the film cassette package may be extended, with suitable calibration, to general diagnostic X-ray survey measurements. (author)

  16. Parallel beam microradiography of dental hard tissue using synchrotron radiation and X-ray image magnification

    International Nuclear Information System (INIS)

    Takagi, S.; Chow, L.C.; Brown, W.E.; Dobbyn, R.C.; Kuriyama, M.

    1984-01-01

    A novel technique utilizing a highly parallel beam of monochromatic synchrotron radiation combined with X-ray image magnification has been used to obtain microradiographs of caries lesions in relatively thick tooth sections. Preliminary results reveal structural features not previously reported. This technique holds the promise of allowing one to follow the structural changes accompanying the formation, destruction and chemical repair of mineralized tissue in real time. (orig.)

  17. Genetic damage in exfoliated cells from oral mucosa of individuals exposed to X-rays during panoramic dental radiographies.

    Science.gov (United States)

    Cerqueira, E M M; Gomes-Filho, I S; Trindade, S; Lopes, M A; Passos, J S; Machado-Santelli, G M

    2004-08-08

    The genotoxic effects of X-ray emitted during dental panoramic radiography were evaluated in exfoliated cells from oral epithelium through a differentiated protocol of the micronucleus test. Thirty-one healthy individuals agreed to participate in this study and were submitted to this procedure for diagnosis purpose after being requested by the dentist. All of them answered a questionnaire before the examination. Cells were obtained from both sides of the cheek by gentle scrapping with a cervical brush, immediately before the exposure and after 10 days. Cytological preparations were stained according to Feulgen-Rossenbeck reaction and analyzed under light and laser scanning confocal microscopies. Micronuclei, nuclear projections (buds and broken eggs) and degenerative nuclear alterations (condensed chromatin, karyolysis and karyorrhexis) were scored. The frequencies of micronuclei, karyolysis and pycnosis were similar before and after exposure (P > 0.90), whereas the condensation of the chromatin and the karyorrhexis increased significantly after exposure (P structures are associated to the normal epithelium differentiation. The results suggest that the X-ray exposure during panoramic dental radiography induces a cytotoxic effect by increasing apoptosis. We also believe that the score of other nuclear alterations in addition to the micronucleus improves the sensitivity of genotoxic effects detection.

  18. A comparison of Kodak Ultraspeed and Ektaspeed Plus dental X-ray films for use in endodontics.

    Science.gov (United States)

    Moule, A J; Wong, A; Monsour, P A; Basford, K E

    2001-06-01

    The advantage of using a faster film for length determination in endodontic therapy is obvious. However, for such a film to be generally accepted, it must demonstrate comparable diagnostic quality to traditionally used films. The comparative accuracy of canal length determination of Ultraspeed and Ektaspeed Plus dental X-ray films was assessed in maxillary first and second molars; for different canals, for different teeth, for different exposures, and for different examiners (five general dentists and three endodontic specialists). In general, there were no significant differences between films, among examiners, or any interaction between films and exposures. That is, an assessor's ability to estimate lengths was not significantly influenced by the film type or by exposure used. There was a wide divergence in the individual assessor's ability to estimate lengths. Specialists estimated lengths more accurately than general practitioners and estimated lengths more accurately with Ektaspeed Plus film. Length determination in distobuccal and mesiobuccal canals was more accurate than in palatal canals. Most palatal canals were underestimated in length by more than 1mm. The use of file sizes larger in number than size 15 is recommended in these canals. For length determination, Ektaspeed Plus dental X-ray film is as effective as Ultraspeed film. Given the acceptable quality and accuracy of Ektaspeed Plus film, there seems to be no clinical reason to subject patients to greater radiation by using a slower film during endodontic therapy.

  19. Free focus radiography with miniaturized dental x-ray machines: a comparison of ''midline'' and ''lateral'' techniques

    International Nuclear Information System (INIS)

    Jensen, T.W.

    1983-01-01

    The use of free focus radiography (FFR) employing miniaturized dental x-ray machines with radiation probes has never been generally accepted in dentistry despite its recognized radiographic potential. The present investigation studied ways to improve imaging and lower radiation burdens in dental free focus radiography. Relatively high air exposures ranging from 42,050 mR per film for high-resolution images to 3,214 mR per film for lower-resolution images using a current midline radiographic technique for panoramic FFR were found. In a proposed lateral FFR panoramic technique, reduced exposures ranged from 420 mR per film for high-resolution images to 14 mR per film for lower-resolution images. In each technique the lower exposure was obtained with a rare earth imaging system. A proposed modification of the current midline FFR technique using a rare earth imaging system and heavy added copper filtration was found to produce exposures in the range normally used in dentistry (207 mr), and the resultant image was high in contrast with relatively low detail. A comparison of essential characteristics of midline and lateral FFR techniques failed to identify specific advantages for the midline technique in current use. Lateral exposure modes in dental FFR should receive increased attention in the interest of good imaging and radiation control. It was noted that existing miniaturized dental x-ray machines may have been designed specifically for use of the midline FFR exposure technique, and modification of this equipment to support reliable lateral exposure modes was recommended

  20. 30 CFR 57.12033 - Hand-held electric tools.

    Science.gov (United States)

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Hand-held electric tools. 57.12033 Section 57.12033 Mineral Resources MINE SAFETY AND HEALTH ADMINISTRATION, DEPARTMENT OF LABOR METAL AND NONMETAL... Surface and Underground § 57.12033 Hand-held electric tools. Hand-held electric tools shall not be...

  1. 30 CFR 56.12033 - Hand-held electric tools.

    Science.gov (United States)

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Hand-held electric tools. 56.12033 Section 56.12033 Mineral Resources MINE SAFETY AND HEALTH ADMINISTRATION, DEPARTMENT OF LABOR METAL AND NONMETAL....12033 Hand-held electric tools. Hand-held electric tools shall not be operated at high potential...

  2. Panoramic dental radiography with an intra-oral X-ray tube

    International Nuclear Information System (INIS)

    Hollender, L.

    1980-01-01

    Surveys of the complete dentition and jaws can be obtained in various ways. Panoramic radiography with an intra-oral X-ray tube results in a lower radiation exposure than other methods, and is less time-consuming. However, the symmetrical views generally employed result in relatively poor imaging of the molars. This disadvantage can be overcome by using lateral projections which also cause less discomfort for the patient and, particularly in combination with intensifying screens, a further reduction in radiation exposure. Some possible future trends are suggested. (Auth.)

  3. Implementation of a radiation protection framework for medical and dental X-ray diagnostic services in Minas Gerais/Brazil

    International Nuclear Information System (INIS)

    Silva, Teogenes A. da; Pereira, Elton G.; Nogueira, Maria do S.; Ferreira, Hudson R.; Alonso, Thessa C.; Castro, Jose G.L. de; Andrade, Mauricio C.; Joana, Georgia S.; Oliveira, Mauricio de; Cezar, Adriana C. Z.

    2008-01-01

    The Brazilian Sanitary Vigilance Agency is the regulatory authority for radiation protection and quality control of all practices with X-rays for diagnostic purpose. In 1998, the technical regulation 'Guidelines for Radiation Protection in Medical and Dental Radiodiagnostic' was issued by the government that reflected the most updated policy recommended by the International Basic Safety Standards for Protection against Ionizing Radiation. To accomplish the objective of improving radiation protection conditions in the state of Minas Gerais, the Development Centre of Nuclear Technology (CDTN) and the Superintendence of Sanitary Vigilance (SVS) signed a formal cooperation agreement that included: an accreditation process for radiation protection professionals, a follow-up program of the services provided by those professionals, technical support from CDTN for audits carried out by SVS and training of SVS inspectors. Actions to improve and assure metrological reliability of the radiation measurements and special attention to mammography services were done. This paper provides details and results of the radiation protection framework for X-ray radiodiagnostic services in Minas Gerais; the success of the adopted model suggests that it can be used as a basic model to other regions. (author)

  4. Behavior analysis in consumer affairs: encouraging dental professionals to provide consumers with shielding from unnecessary X-ray exposure

    International Nuclear Information System (INIS)

    Greene, B.F.; Neistat, M.D.

    1983-01-01

    An unobtrusive observation system was developed to determine the extent to which dental professionals in two communities provided lead shielding to patients during X-ray exams. A lengthy baseline revealed low and irregular provision of shielding among half of these professionals. Subsequently, a program was undertaken by a consumer's group in which these professionals were requested to provide shielding and were given confidential feedback regarding its use during the baseline period. The provision of shielding dramatically increased at all offices and was maintained throughout a follow-up period extending to more than 9 months after the program's implementation. Little or no generalized effect was observed in the occurrence of three collateral behaviors that were also assessed throughout the study

  5. Micro energy-dispersive X-ray fluoresence mapping of enamel and dental materials after chemical erosion.

    Science.gov (United States)

    Soares, Luís Eduardo Silva; de Oliveira, Rodrigo; Nahórny, Sídnei; Santo, Ana Maria do Espírito; Martin, Airton Abrahão

    2012-10-01

    Energy-dispersive X-ray fluorescence was employed to test the hypothesis that beverage consumption or mouthwash utilization will change the chemical properties of dental materials and enamel mineral content. Bovine enamel samples (n = 45) each received two cavity preparations (n = 90), each pair filled with one of three dental materials (R: nanofilled composite resin; GIC: glass-ionomer cement; RMGIC: resin-modified GIC). Furthermore, they were treated with three different solutions (S: saliva; E: erosion/Pepsi Twist®; or EM: erosion+mouthwash/Colgate Plax®). It was found that mineral loss in enamel was greater in GICE samples than in RE > RMGICE > RMGICEM > REM > GICEM. An increased percentage of Zr was found in REM indicating organic matrix degradation. Dental materials tested (R, GIC, and RMGIC) were not able to protect adjacent enamel from acid erosion by the soft drink tested. The use of mouthwash promoted protection of enamel after erosion by the soft drink. To avoid chemical dissolution by mouthwashes, protection by resin composites with surface sealants is recommended.

  6. Hand-held optical fuel pin scanner

    International Nuclear Information System (INIS)

    Kirchner, T.L.; Powers, H.G.

    1987-01-01

    A portable, hand-held apparatus is described for optically scanning indicia imprinted about a planar end face of an article having an outer wall surface, the apparatus comprising: a supporting frame; light detector means fixed to the frame for digitizing light patterns directed thereto; indexing means on the frame for engaging the planar end face and locating the end face in a preselected focal plane on the frame. The indexing means has an inner wall surface complementary to the article wall surface for disposition thereabout and terminates in an end portion beyond the planar end face. The inner wall surface has a radially inwardly extending shoulder spaced from the end portion and engageable with the planar end face; light means directed onto the preselected focal plane; optical means mounted on the frame about a central axis, the optical means being optically interposed between the indexing means and the light detector means for directing reflected light from the preselected focal plane to the light detector means and including a dove prism centrally aligned along the central axis; and means for selectively rotating the dove prism relative to the frame about the central axis to thereby rotate the image from the focal plane as transmitted to the light detector means

  7. Gen-2 Hand-Held Optical Imager towards Cancer Imaging: Reflectance and Transillumination Phantom Studies

    Directory of Open Access Journals (Sweden)

    Anuradha Godavarty

    2012-02-01

    Full Text Available Hand-held near-infrared (NIR optical imagers are developed by various researchers towards non-invasive clinical breast imaging. Unlike these existing imagers that can perform only reflectance imaging, a generation-2 (Gen-2 hand-held optical imager has been recently developed to perform both reflectance and transillumination imaging. The unique forked design of the hand-held probe head(s allows for reflectance imaging (as in ultrasound and transillumination or compressed imaging (as in X-ray mammography. Phantom studies were performed to demonstrate two-dimensional (2D target detection via reflectance and transillumination imaging at various target depths (1–5 cm deep and using simultaneous multiple point illumination approach. It was observed that 0.45 cc targets were detected up to 5 cm deep during transillumination, but limited to 2.5 cm deep during reflectance imaging. Additionally, implementing appropriate data post-processing techniques along with a polynomial fitting approach, to plot 2D surface contours of the detected signal, yields distinct target detectability and localization. The ability of the gen-2 imager to perform both reflectance and transillumination imaging allows its direct comparison to ultrasound and X-ray mammography results, respectively, in future clinical breast imaging studies.

  8. X-ray crystallography

    Science.gov (United States)

    2001-01-01

    X-rays diffracted from a well-ordered protein crystal create sharp patterns of scattered light on film. A computer can use these patterns to generate a model of a protein molecule. To analyze the selected crystal, an X-ray crystallographer shines X-rays through the crystal. Unlike a single dental X-ray, which produces a shadow image of a tooth, these X-rays have to be taken many times from different angles to produce a pattern from the scattered light, a map of the intensity of the X-rays after they diffract through the crystal. The X-rays bounce off the electron clouds that form the outer structure of each atom. A flawed crystal will yield a blurry pattern; a well-ordered protein crystal yields a series of sharp diffraction patterns. From these patterns, researchers build an electron density map. With powerful computers and a lot of calculations, scientists can use the electron density patterns to determine the structure of the protein and make a computer-generated model of the structure. The models let researchers improve their understanding of how the protein functions. They also allow scientists to look for receptor sites and active areas that control a protein's function and role in the progress of diseases. From there, pharmaceutical researchers can design molecules that fit the active site, much like a key and lock, so that the protein is locked without affecting the rest of the body. This is called structure-based drug design.

  9. The use of the X-ray investigation for dental caries diagnostics

    International Nuclear Information System (INIS)

    Cecetkova, A.; Ondrasovicova, J.

    2008-01-01

    Caries is defect of the hard dental tissue which is cause by cariogenic microorganisms in the oral cavity. New classification of caries is D1, D2, D3 and D4. D1 is caries in the first layer of enamel. D2 is caries in the second layer of enamel till the margin of dentine ( caries superficial ). D3 is caries in the first layer of dentine ( caries media ).D4 is caries on the top of pulp chamber ( caries pulpae proxima ). One of the most frequently used methods of radiography to detect of caries is bitewing ( BTW ) radiography. (authors)

  10. MONITORING OF INDIVIDUAL DOSES FOR MEDICAL WORKERS OF DENTAL POLYCLINIC’S X-RAY ROOMS IN DUSHANBE, THE REPUBLIC OF TADJIKISTAN

    Directory of Open Access Journals (Sweden)

    N. U. Hakimova

    2016-01-01

    Full Text Available The article presents the data and analyses of personnel’s average annual external exposure doses monitoring via the thermoluminescent dosimetry method used for X-ray radiological personnel in dental polyclinics of Dushanbe, Tadjikistan Republic over a 5-year period ( 2010–2014 . Out of 42 registered medical institutions dental polyclinics amounted up to only just 14%. For this work thermoluminescent dosimeters were used ( with LiF: Mg, Ti with the thermoluminescent dosimetric installation “ Harshaw – 4500” as the reader device. Monitoring results comparison of individual dose equivalent Hp ( 10 values was conducted for two groups of medical workers: medical doctors and X-ray lab technicians. It is demonstrated that radiological technicians’ professional exposure doses are on the average by 23% higher than those for medical doctors.The average individual exposure doses over the above indicated period amount to 0,93 mSv and 1,3 mSv for doctors and X-ray lab technicians, respectively, and are in the range from 0,45 mSv to 2,39 mSv. The doses include contribution from the natural background. The values of doses recorded for the personnel in dental polyclinic correspond to those recorded for the workers in the routine X-ray rooms.

  11. Some measurements of doses to patients from dental X-rays

    International Nuclear Information System (INIS)

    Woehni, T.

    1976-01-01

    Some measurements of doses to patients from conventional dental radiography and orthopantomography are presented. Doses to the red bone marrow are calculated. The bone marrow doses from two different exposures, Maxilla incisor and Molar bite-wing, were calculated to be 0.4 and 1.0 mrad respectively. The average dose to red bone marrow from a full-mouth examination (10 exposures) was 0.7 mrad/exposure. An orthopantomographic examination involved 2 mrad to the bone marrow. The greatest doses from an orthopantomographic examination were found around the lateral rotational axis, namely 700 mrad. The dose distributions from the two different cone lengths did not differ as much as expected, mainly due to scattered radiation. (Auth.)

  12. Some measurements of doses to patients from dental X-rays

    Energy Technology Data Exchange (ETDEWEB)

    Woehni, T [Statens Institutt for Straalehygiene, Oslo (Norway)

    1976-11-01

    Some measurements of doses to patients from conventional dental radiography and orthopantomography are presented. Doses to the red bone marrow are calculated. The bone marrow doses from two different exposures, Maxilla incisor and Molar bite-wing, were calculated to be 0.4 and 1.0 mrad respectively. The average dose to red bone marrow from a full-mouth examination (10 exposures) was 0.7 mrad/exposure. An orthopantomographic examination involved 2 mrad to the bone marrow. The greatest doses from an orthopantomographic examination were found around the lateral rotational axis, namely 700 mrad. The dose distributions from the two different cone lengths did not differ as much as expected, mainly due to scattered radiation.

  13. Characterization of Scattered X-Ray Photons in Dental Cone-Beam Computed Tomography.

    Science.gov (United States)

    Yang, Ching-Ching

    2016-01-01

    Scatter is a very important artifact causing factor in dental cone-beam CT (CBCT), which has a major influence on the detectability of details within images. This work aimed to improve the image quality of dental CBCT through scatter correction. Scatter was estimated in the projection domain from the low frequency component of the difference between the raw CBCT projection and the projection obtained by extrapolating the model fitted to the raw projections acquired with 2 different sizes of axial field-of-view (FOV). The function for curve fitting was optimized by using Monte Carlo simulation. To validate the proposed method, an anthropomorphic phantom and a water-filled cylindrical phantom with rod inserts simulating different tissue materials were scanned using 120 kVp, 5 mA and 9-second scanning time covering an axial FOV of 4 cm and 13 cm. The detectability of the CT image was evaluated by calculating the contrast-to-noise ratio (CNR). Beam hardening and cupping artifacts were observed in CBCT images without scatter correction, especially in those acquired with 13 cm FOV. These artifacts were reduced in CBCT images corrected by the proposed method, demonstrating its efficacy on scatter correction. After scatter correction, the image quality of CBCT was improved in terms of target detectability which was quantified as the CNR for rod inserts in the cylindrical phantom. Hopefully the calculations performed in this work can provide a route to reach a high level of diagnostic image quality for CBCT imaging used in oral and maxillofacial structures whilst ensuring patient dose as low as reasonably achievable, which may ultimately make CBCT scan a reliable and safe tool in clinical practice.

  14. Characterization of Scattered X-Ray Photons in Dental Cone-Beam Computed Tomography.

    Directory of Open Access Journals (Sweden)

    Ching-Ching Yang

    Full Text Available Scatter is a very important artifact causing factor in dental cone-beam CT (CBCT, which has a major influence on the detectability of details within images. This work aimed to improve the image quality of dental CBCT through scatter correction.Scatter was estimated in the projection domain from the low frequency component of the difference between the raw CBCT projection and the projection obtained by extrapolating the model fitted to the raw projections acquired with 2 different sizes of axial field-of-view (FOV. The function for curve fitting was optimized by using Monte Carlo simulation. To validate the proposed method, an anthropomorphic phantom and a water-filled cylindrical phantom with rod inserts simulating different tissue materials were scanned using 120 kVp, 5 mA and 9-second scanning time covering an axial FOV of 4 cm and 13 cm. The detectability of the CT image was evaluated by calculating the contrast-to-noise ratio (CNR.Beam hardening and cupping artifacts were observed in CBCT images without scatter correction, especially in those acquired with 13 cm FOV. These artifacts were reduced in CBCT images corrected by the proposed method, demonstrating its efficacy on scatter correction. After scatter correction, the image quality of CBCT was improved in terms of target detectability which was quantified as the CNR for rod inserts in the cylindrical phantom.Hopefully the calculations performed in this work can provide a route to reach a high level of diagnostic image quality for CBCT imaging used in oral and maxillofacial structures whilst ensuring patient dose as low as reasonably achievable, which may ultimately make CBCT scan a reliable and safe tool in clinical practice.

  15. Guideline development and impact assessment for registration of medical, dental and veterinary x-ray apparatus

    International Nuclear Information System (INIS)

    Colgan, P.; Harrison, D.; Moore, W.

    1996-01-01

    Under the NSW Radiation Control Act 1990, radiation apparatus used for diagnostic medical, dental and veterinary purposes will be required to become registered. The inspection required prior to registration will be conducted by a Consulting Radiation Expert who has been accredited by the Environment Protection Authority (EPA) as being competent in the field of quality assurance assessment of radiation apparatus used for diagnostic medical, dental and veterinary purposes. When regulating any activity in NSW, there is a requirement to undertake a regulatory impact statement of the proposed regulation. In addition, the introduction of any accompanying guideline requires a cost-benefit analysis. Costs may include enforcement, administrative and compliance activities. The calculation of benefit relies heavily on the improvement in apparatus performance (and hence dose reduction) that can be obtained with the introduction of a mandatory practice such as apparatus registration. This paper discusses the development of the registration guideline for NSW, including a summary of the public comments received. It further discusses the methodology and data used for the accompanying cost-benefit analysis. Information in this paper is presented in three parts: EPA field survey, cost analysis, and benefit analysis. For NSW it was estimated that the introduction of registration of these apparatus, over a two year period, would result in early replacement and repair costs (present values) to the medical industry of between $5.7 and $11.0 million, with an additional $2.5 million in EPA enforcement costs. The introduction of the proposed system of registration is expected to result in an estimated savings in quantifiable health detriment costs to NSW of between $11.8 and $17.7 million, and reduce the risk of radiation induced mortality. (authors)

  16. Analysis of Dental Enamel Surface Submitted to Fruit Juice Plus Soymilk by Micro X-Ray Fluorescence: In Vitro Study

    Directory of Open Access Journals (Sweden)

    Janaína Salmos Brito

    2016-01-01

    Full Text Available Objective. This paper aimed to analyze the in vitro industrialized fruit juices effect plus soy to establish the erosive potential of these solutions. Materials and Methods. Seventy bovine incisors were selected after being evaluated under stereomicroscope. Their crowns were prepared and randomly divided into 7 groups, using microhardness with allocation criteria. The crowns were submitted to the fruit juice plus soy during 15 days, twice a day. The pH values, acid titration, and Knoop microhardness were recorded and the specimens were evaluated using X-ray microfluorescence (µXRF. Results. The pH average for all juices and after 3 days was significantly below the critical value for dental erosion. In average, the pH value decreases 14% comparing initial time and pH after 3 days. Comparing before and after, there was a 49% microhardness decrease measured in groups (p<0.05. Groups G1, G2, G5, and G6 are above this average. The analysis by μXRF showed a decrease of approximately 7% Ca and 4% P on bovine crowns surface. Florida (FL statistical analysis showed a statistically significant 1 difference between groups. Thus, a tooth chance to suffer demineralization due to industrialized fruit juices plus soy is real.

  17. A comparison of density of Insight and Ektaspeed plus dental x-ray films using automatic and manual processing

    International Nuclear Information System (INIS)

    Yoon, Suk Ja

    2001-01-01

    To compare the film density of Insight dental X-ray film (Eastman Kodak Co., Rochester, NY, USA) with that of Ektaspeed Plus film (Eastman Kodak) under manual and automatic processing conditions. Insight and wedge on the film under the three different exposure times. The exposed films were processed by both manual and automatic ways. The Base plus fog density and the optical density and the optical density made by exposing step wedge were calculated using a digital densitometer (model 07-443, Victoreen Inc, Cleveland, Ohio, USA). The optical densities of the Insight and Ektaspeed film versus thickness of aluminum wedge at the same exposure time were plotted on the graphs. Statistical analyses were applied for comparing the optical densities of the two films. The film density of both Insight films and Ektaspeed Plus films under automatic processing condition was significantly higher over the manual processing. The film density of Insight over Ektaspeed Plus film. To take the full advantage of reducing exposure time, Insight film should be processed automatically

  18. Unique migration of a dental needle into the parapharyngeal space: successful removal by an intraoral approach and simulation for tracking visibility in X-ray fluoroscopy.

    Science.gov (United States)

    Okumura, Yuri; Hidaka, Hiroshi; Seiji, Kazumasa; Nomura, Kazuhiro; Takata, Yusuke; Suzuki, Takahiro; Katori, Yukio

    2015-02-01

    The first objective was to describe a novel case of migration of a broken dental needle into the parapharyngeal space. The second was to address the importance of simulation elucidating visualization of such a thin needle under X-ray fluoroscopy. Clinical case records (including computed tomography [CT] and surgical approaches) were reviewed, and a simulation experiment using a head phantom was conducted using the same settings applied intraoperatively. A 36-year-old man was referred after failure to locate a broken 31-G dental needle. Computed tomography revealed migration of the needle into the parapharyngeal space. Intraoperative X-ray fluoroscopy failed to identify the needle, so a steel wire was applied as a reference during X-ray to locate the foreign body. The needle was successfully removed using an intraoral approach with tonsillectomy under surgical microscopy. The simulation showed that the dental needle was able to be identified only after applying an appropriate compensating filter, contrasting with the steel wire. Meticulous preoperative simulation regarding visual identification of dental needle foreign bodies is mandatory. Intraoperative radiography and an intraoral approach with tonsillectomy under surgical microscopy offer benefits for accessing the parapharyngeal space, specifically for cases medial to the great vessels. © The Author(s) 2014.

  19. Investigation of the structure of human dental tissue at multiple length scales using high energy synchrotron X-ray SAXS/WAXS

    Science.gov (United States)

    Sui, Tan; Landini, Gabriel; Korsunsky, Alexander M.

    2011-10-01

    High energy (>50keV) synchrotron X-ray scattering experiments were carried out on beamline I12 JEEP at the Diamond Light Source (DLS, Oxford, UK). Although a complete human tooth could be studied, in the present study attention was focused on coupons from the region of the Dentin-Enamel Junction (DEJ). Simultaneous high energy SAXS/WAXS measurements were carried out. Quantitative analysis of the results allows multiple length scale characterization of the nano-crystalline structure of dental tissues. SAXS patterns analysis provide insight into the mean thickness and orientation of hydroxyapatite particles, while WAXS (XRD) patterns allow the determination of the crystallographic unit cell parameters of the hydroxyapatite phase. It was found that the average particle thickness determined from SAXS interpretation varies as a function of position in the vicinity of the DEJ. Most mineral particles are randomly orientated within dentin, although preferred orientation emerges and becomes stronger on approach to the enamel. Within the enamel, texture is stronger than anywhere in the dentin, and the determination of lattice parameters can be accomplished by Pawley refinement of the multiple peak diffraction pattern. The results demonstrate the feasibility of using high energy synchrotron X-ray beams for the characterization of human dental tissues. This opens up the opportunity of studying thick samples (e.g., complete teeth) in complex sample environments (e.g., under saline solution). This opens new avenues for the application of high energy synchrotron X-ray scattering to dental research.

  20. The availability of relatively cheap hand-held Global Positioning ...

    African Journals Online (AJOL)

    spamer

    conditions, so the approach failed to produce results ... Hand-held Global Positioning System (GPS) receivers provide opportunities for detailed and rapid mapping of features ..... TICKELL, W. L. N. 1968 — The biology of the great albatrosses,.

  1. An Intelligent Hand-Held Microsurgical Instrument for Improved Accuracy

    National Research Council Canada - National Science Library

    Ang, Wei

    2001-01-01

    This paper presents the development and initial experimental results of the first prototype of Micron, an active hand-held instrument to sense and compensate physiological tremor and other unwanted...

  2. Plans for Hand-Held/Portable Oil Assessment Devices

    National Research Council Canada - National Science Library

    Urbansky, Edward

    2005-01-01

    At the request of the U.S. Army Oil Analysis Program, the JOAP TSC conducted a market study, assembled a plan of action, and prepared a worksheet for the evaluation of portable or hand-held oil assessment devices...

  3. Dental x-rays

    Science.gov (United States)

    ... Aps JK. Radiographic techniques. In: Dean JA, ed. McDonald and Avery's Dentistry for the Child and Adolescent . ... M. is also a founding member of Hi-Ethics and subscribes to the principles of the Health ...

  4. Drywall plates evaluation as protection barriers in dental X-rays and mammography facilities; Evaluacion de placas de drywall como barreras de proteccion en instalaciones de rayos X dental y mamografia

    Energy Technology Data Exchange (ETDEWEB)

    Guevara R, V. Y.; Romero C, N. [Empresa QC DOSE S. A. C., Av. Tomas Marsano 1915, Surquillo, Lima 34 (Peru); Berrocal T, M., E-mail: vguevara@qcdose.com [Universidad Nacional Mayor de San Marcos, C. German Amezaga 375, Edif. Jorge Basadre, Ciudad Universitaria, Lima 1 (Peru)

    2014-08-15

    In the dental X-rays and mammography facilities, usually lead shielded walls as protective barriers are used. Lead is a good attenuator for X-rays, but has toxic properties and its cost is high. Mammography equipment s emit low-energy photons in the range of 25 keV to 35 keV, on current computers; the primary radiation beam is intercepted by the image receptor. Peri apical dental equipment s emit photons in the range of 50 to 90 keV, their filtration is fixed. These devices emit a collimated beam whose diameter is slightly larger than the diagonal dimension of a standard film of dental X-rays. When a dental x-ray is performed, the radiation beam is partially attenuated by the patient. Drywall is a material consisting of plasterboard between two cardboard layers, being its components gypsum and cellulose generally. It is used in construction for execution of interior walls, ceilings and wall coverings, could also serve as a replacement for lead as well as other materials. In this paper three drywall prototypes (Giplac), formed with 02, 04 and 06 drywall layers (13, 16 and 20 cm of thickness respectively) were tested as barriers against primary and secondary X-ray radiation that come from dental and mammography equipment s. The results show that the drywall prototype, 02 layers, efficiently attenuates the secondary radiation beam produced by conventional mammography equipment. And the prototype 04 and 06 layers, efficiently attenuates the primary radiation beam produced by peri apical dental equipment. (author)

  5. Skull x-ray

    Science.gov (United States)

    X-ray - head; X-ray - skull; Skull radiography; Head x-ray ... There is low radiation exposure. X-rays are monitored and regulated to provide the minimum amount of radiation exposure needed to produce the image. Most ...

  6. Neck x-ray

    Science.gov (United States)

    X-ray - neck; Cervical spine x-ray; Lateral neck x-ray ... There is low radiation exposure. X-rays are monitored so that the lowest amount of radiation is used to produce the image. Pregnant women and ...

  7. Microstructural characterization of dental zinc phosphate cements using combined small angle neutron scattering and microfocus X-ray computed tomography

    Czech Academy of Sciences Publication Activity Database

    Viani, Alberto; Sotiriadis, Konstantinos; Kumpová, Ivana; Mancini, L.; Appavou, M.-S.

    2017-01-01

    Roč. 33, č. 4 (2017), s. 402-417 ISSN 0109-5641 R&D Projects: GA MŠk(CZ) LO1219 Keywords : zinc phosphate cements * small angle neutron scattering * X-ray micro-computed tomography * X-ray powder diffraction * zinc oxide * acid-base cements Subject RIV: JJ - Other Materials OBOR OECD: Composites (including laminates, reinforced plastics, cermets, combined natural and synthetic fibre fabrics Impact factor: 4.070, year: 2016 https://www.sciencedirect.com/science/article/pii/S0109564116305127

  8. Performance of a gaseous detector based energy dispersive X-ray fluorescence imaging system: Analysis of human teeth treated with dental amalgam

    International Nuclear Information System (INIS)

    Silva, A.L.M.; Figueroa, R.; Jaramillo, A.; Carvalho, M.L.; Veloso, J.F.C.A.

    2013-01-01

    Energy dispersive X-ray fluorescence (EDXRF) imaging systems are of great interest in many applications of different areas, once they allow us to get images of the spatial elemental distribution in the samples. The detector system used in this study is based on a micro patterned gas detector, named Micro-Hole and Strip Plate. The full field of view system, with an active area of 28 × 28 mm 2 presents some important features for EDXRF imaging applications, such as a position resolution below 125 μm, an intrinsic energy resolution of about 14% full width at half maximum for 5.9 keV X-rays, and a counting rate capability of 0.5 MHz. In this work, analysis of human teeth treated by dental amalgam was performed by using the EDXRF imaging system mentioned above. The goal of the analysis is to evaluate the system capabilities in the biomedical field by measuring the drift of the major constituents of a dental amalgam, Zn and Hg, throughout the tooth structures. The elemental distribution pattern of these elements obtained during the analysis suggests diffusion of these elements from the amalgam to teeth tissues. - Highlights: • Demonstration of an EDXRF imaging system based on a 2D-MHSP detector for biological analysis • Evaluation of the drift of the dental amalgam constituents, throughout the teeth • Observation of Hg diffusion, due to hydroxyapatite crystal defects that compose the teeth tissues

  9. Performance of a gaseous detector based energy dispersive X-ray fluorescence imaging system: Analysis of human teeth treated with dental amalgam

    Energy Technology Data Exchange (ETDEWEB)

    Silva, A.L.M. [I3N, Physics Dept, University of Aveiro, 3810-193 Aveiro (Portugal); Figueroa, R.; Jaramillo, A. [Physics Department, Universidad de La Frontera, Temuco (Chile); Carvalho, M.L. [Atomic Physics Centre, University of Lisbon, 1649-03 Lisboa (Portugal); Veloso, J.F.C.A., E-mail: joao.veloso@ua.pt [I3N, Physics Dept, University of Aveiro, 3810-193 Aveiro (Portugal)

    2013-08-01

    Energy dispersive X-ray fluorescence (EDXRF) imaging systems are of great interest in many applications of different areas, once they allow us to get images of the spatial elemental distribution in the samples. The detector system used in this study is based on a micro patterned gas detector, named Micro-Hole and Strip Plate. The full field of view system, with an active area of 28 × 28 mm{sup 2} presents some important features for EDXRF imaging applications, such as a position resolution below 125 μm, an intrinsic energy resolution of about 14% full width at half maximum for 5.9 keV X-rays, and a counting rate capability of 0.5 MHz. In this work, analysis of human teeth treated by dental amalgam was performed by using the EDXRF imaging system mentioned above. The goal of the analysis is to evaluate the system capabilities in the biomedical field by measuring the drift of the major constituents of a dental amalgam, Zn and Hg, throughout the tooth structures. The elemental distribution pattern of these elements obtained during the analysis suggests diffusion of these elements from the amalgam to teeth tissues. - Highlights: • Demonstration of an EDXRF imaging system based on a 2D-MHSP detector for biological analysis • Evaluation of the drift of the dental amalgam constituents, throughout the teeth • Observation of Hg diffusion, due to hydroxyapatite crystal defects that compose the teeth tissues.

  10. Radiation exposure with the NOMAD portable X-ray system.

    Science.gov (United States)

    Goren, A D; Bonvento, M; Biernacki, J; Colosi, D C

    2008-02-01

    A new hand-held battery-operated portable X-ray system was tested for possible leakage radiation through the existing heavy metal compounds surrounding the X-ray tube, backscatter radiation through the lead-filled acrylic shield attached at the end of the exit tube and patient exposure. Dose measurements were conducted using a DXTRR phantom and a water phantom. All measurements were recorded using calibrated thermoluminescent dosimetry (TLD), calibrated Unfors Model 583L dosemeter, and a calibrated Radcal MDH model 1015 dosemeter. The settings for all exposure were 60 kVp, 2.3 mA and 0.25 s using Kodak Insight (Class F) film. All backscatter measurements, in front of the shield, behind the shield, at the finger of the operator, the operator's chest, eyes and gonads were significantly below the maximum permissible radiation leakage as per the United States Food and Drug Administration regulations (100 mR h(-1)). Our measurements indicate that the exposure would be well within the occupational maximum permissible dose for an occupationally exposed person. Film dose was consistent with the manufacturer's recommendations. As a result of our measurements, the State of New York Bureau of Environmental Radiation Protection granted us a variance to use the NOMAD on a case-by-case basis. Our data have shown that the NOMAD presents risks that are no greater than with standard dental radiographic units to the patient or operator and the measured doses are well below recommended levels.

  11. Epilepsy Forewarning Using A Hand-Held Device

    Energy Technology Data Exchange (ETDEWEB)

    Hively, LM

    2005-02-21

    Over the last decade, ORNL has developed and patented a novel approach for forewarning of a large variety of machine and biomedical events. The present implementation uses desktop computers to analyze archival data. This report describes the next logical step in this effort, namely use of a hand-held device for the analysis.

  12. Development of a concept for radiation patients exposure assessment during dental X-ray examinations and statistical data acquisition for the determination of a diagnostic reference value; Erarbeitung eines Konzepts zur Ermittlung der Strahlenexposition von Patienten bei zahnmedizinischen Roentgenuntersuchungen und Erhebung von statistischen Daten zur Erstellung diagnostischer Referenzwerten

    Energy Technology Data Exchange (ETDEWEB)

    Kueppers, C.; Sering, M. [Oekoinstitut Freiburg e.V., Freiburg (Germany); Poppe, B.; Poplawski, A.; Looe, H.K.; Beyer, D.; Pfaffenberger, A.; Chofor, N.; Eenboom, F. [Pius Hospital, Oldenburg (Germany)

    2012-02-15

    The research project on the development a concept for radiation patients exposure assessment during dental X-ray examinations and statistical data acquisition for the determination of a diagnostic reference value includes the following issues: Fundamental facts: dental X-ray examination techniques, dose relevant factors and characteristics during X-ray examinations, radiation exposed organs during dental X-ray examinations, dose assessment based on phantoms. Materials and methodologies of the project: TLD measurements using the phantom, calculation of the effective dose during dental X-ray examinations, properties and settings of the reference facilities for the determination of radiation exposure, selection of dental offices, dosimetric measurements, data acquisition and statistical evaluation. Results of dosimetric examinations: results of dosimetric measurements at reference facilities, results of dosimetric measurements in dental offices. Discussion of the concept for the determination of the radiation exposure during dental X-ray examinations.

  13. TU-FG-209-09: Mathematical Estimation and Experimental Measurement of Patient Free-In-Air Skin Entrance Exposure During a Panoramic Dental X-Ray Procedure

    Energy Technology Data Exchange (ETDEWEB)

    Errico, A; Behrman, R; Li, B [Boston University Medical Center, Boston, MA (United States)

    2016-06-15

    Purpose: To develop a simple mathematical model for estimating the patient free-in-air skin entrance exposure (SEE) during a panoramic dental x-ray that does not require the use of a head phantom. This eliminates issues associated with phantom centering and the mounting of a detector on the phantom for routine QC testing. Methods: We used a Sirona Orthophos XG panoramic radiographic unit and a Radcal Accu-Gold system for this study. A solid state detector was attached over the slit of the Orthophos’ sensor with the help of a custom-built jig. A single measurement of the free-in-air exposure at this position was taken over a full panoramic scan. A mathematical model for estimating the SEE was developed based upon this measurement, the system geometry, x-ray field beam width, and x-ray sweep angle. To validate the model, patient geometry was simulated by a 16 cm diameter PMMA CTDI phantom centered at the machine’s isocenter. Measurements taken on the phantom’s surface were made using a solid state detector with lead backing, an ion chamber, and the ion chamber with the phantom wrapped in lead to mitigate backscatter. Measurements were taken near the start position of the tube and at 90 degrees from the start position. Results: Using the solid state detector, the average SEE was 23.5+/−0.02 mR and 55.5+/−0.08 mR at 64 kVp and 73 kVp, respectively. With the lead-wrapping, the measurements from the ion chamber matched those of the solid state detector to within 0.1%. Preliminary results gave the difference between the mathematical model and the phantom measurements to be approximately 5% at both kVps. Conclusion: Reasonable estimates of patient SEE for panoramic dental radiography can be made using a simple mathematical model without the need for a head phantom.

  14. TU-FG-209-09: Mathematical Estimation and Experimental Measurement of Patient Free-In-Air Skin Entrance Exposure During a Panoramic Dental X-Ray Procedure

    International Nuclear Information System (INIS)

    Errico, A; Behrman, R; Li, B

    2016-01-01

    Purpose: To develop a simple mathematical model for estimating the patient free-in-air skin entrance exposure (SEE) during a panoramic dental x-ray that does not require the use of a head phantom. This eliminates issues associated with phantom centering and the mounting of a detector on the phantom for routine QC testing. Methods: We used a Sirona Orthophos XG panoramic radiographic unit and a Radcal Accu-Gold system for this study. A solid state detector was attached over the slit of the Orthophos’ sensor with the help of a custom-built jig. A single measurement of the free-in-air exposure at this position was taken over a full panoramic scan. A mathematical model for estimating the SEE was developed based upon this measurement, the system geometry, x-ray field beam width, and x-ray sweep angle. To validate the model, patient geometry was simulated by a 16 cm diameter PMMA CTDI phantom centered at the machine’s isocenter. Measurements taken on the phantom’s surface were made using a solid state detector with lead backing, an ion chamber, and the ion chamber with the phantom wrapped in lead to mitigate backscatter. Measurements were taken near the start position of the tube and at 90 degrees from the start position. Results: Using the solid state detector, the average SEE was 23.5+/−0.02 mR and 55.5+/−0.08 mR at 64 kVp and 73 kVp, respectively. With the lead-wrapping, the measurements from the ion chamber matched those of the solid state detector to within 0.1%. Preliminary results gave the difference between the mathematical model and the phantom measurements to be approximately 5% at both kVps. Conclusion: Reasonable estimates of patient SEE for panoramic dental radiography can be made using a simple mathematical model without the need for a head phantom.

  15. A statistical survey of x-ray CT cases at the Clinic of Oral and Maxillofacial Radiology, Chiba Hospital, Tokyo Dental College

    International Nuclear Information System (INIS)

    Kitagawa, Hiromi; Wakoh, Mamoru; Shibuya, Hitoshi; Yamada, Masayuki; Harada, Takuya; Yamamoto, Kazuhiro; Makihara, Masahiro; Kuroyanagi, Kinya

    1997-01-01

    Statistical study was performed of x-ray CT cases at the Clinic of Oral and Maxillofacial Radiology, Chiba Hospital, Tokyo Dental College since a Toshiba CT Scanner TCT-700S was settled in 1988, until December 1994. Total number of cases photographed was 2645 cases, Male; 1447 (54.7%), Female; 1198 cases (45.3%). Total number of cases yearly photographed have increased every year. 95.43% of the cases were the diseases of oral surgery regions. X-ray CT has been used to malignant tumors (730 cases; 29.7%), cyst (435 cases; 17.7%), inflammation (362 cases; 14.7%), benign tumors (261 cases; 10.6%), injury (171 cases; 7.0%), salivary gland diseases (126 cases; 5.1%) and others. The number of tumors and cyst have been increasing every year. Average number of slices in every diseases were counted. Malignant tumor, injury, temporomandibular joint diseases, and congenital anomalies and malformations were counted many slices. Percentages of number of enhanced CT cases have been increased every year. Recently, number of enhanced CT cases have more number than non-enhanced CT cases. This attitude is correlated with the number of malignant tumors which have been increasing every year. Total number of cases of three dimensional imaging CT (3D-CT) was 316 cases. 3D-CT has been used to injury (146 cases; 46.2%), temporomandibular joint diseases (52 cases; 16.4%), congenital anomalies and malformations (49 cases; 15.5%), tumors (21 cases; 6.7%), cyst (13 cases; 4.1%) and others. The need of x-ray CT in our field and the tendency of dental treatment at Chiba Hospital might be changed in the future. In order to this situation, this type of statistical study will be performed again. (author)

  16. Performance of a gaseous detector based energy dispersive X-ray fluorescence imaging system: Analysis of human teeth treated with dental amalgam

    Science.gov (United States)

    Silva, A. L. M.; Figueroa, R.; Jaramillo, A.; Carvalho, M. L.; Veloso, J. F. C. A.

    2013-08-01

    Energy dispersive X-ray fluorescence (EDXRF) imaging systems are of great interest in many applications of different areas, once they allow us to get images of the spatial elemental distribution in the samples. The detector system used in this study is based on a micro patterned gas detector, named Micro-Hole and Strip Plate. The full field of view system, with an active area of 28 × 28 mm2 presents some important features for EDXRF imaging applications, such as a position resolution below 125 μm, an intrinsic energy resolution of about 14% full width at half maximum for 5.9 keV X-rays, and a counting rate capability of 0.5 MHz. In this work, analysis of human teeth treated by dental amalgam was performed by using the EDXRF imaging system mentioned above. The goal of the analysis is to evaluate the system capabilities in the biomedical field by measuring the drift of the major constituents of a dental amalgam, Zn and Hg, throughout the tooth structures. The elemental distribution pattern of these elements obtained during the analysis suggests diffusion of these elements from the amalgam to teeth tissues.

  17. X-Ray

    Science.gov (United States)

    ... enema. What you can expect During the X-ray X-rays are performed at doctors' offices, dentists' offices, ... as those using a contrast medium. Your child's X-ray Restraints or other techniques may be used to ...

  18. Abdominal x-ray

    Science.gov (United States)

    Abdominal film; X-ray - abdomen; Flat plate; KUB x-ray ... There is low radiation exposure. X-rays are monitored and regulated to provide the minimum amount of radiation exposure needed to produce the image. Most ...

  19. Chest X-Ray

    Medline Plus

    Full Text Available ... talk with you about chest radiography also known as chest x-rays. Chest x-rays are the ... treatment for a variety of lung conditions such as pneumonia, emphysema and cancer. A chest x-ray ...

  20. Hand-held electronic data collection and procedure environment

    International Nuclear Information System (INIS)

    Kennedy, E.; Doniz, K.

    1996-01-01

    As part of a CANDU Owners Group project, AECL has developed a hand-held electronic data collection and procedure environment. Integral to this environment is the C omputerized Procedure Engine . The development of the CPE allows operators, maintainers, and technical support staff to execute virtually any type of station procedure on a general-purpose PC-compatible hand-held computer. There are several advantages to using the computerized procedures: less paper use and handling, reduction in human error, reduction in rework in the field, an increase in procedural compliance, and immediate availability of data to download to databases and plant information systems. The paper describes: the advantages of using computerized procedures, why early forms of computerized procedures were inadequate, the features that the C omputerized Procedure Engine o ffers to the user, the streamlined life cycle of a computerized procedure, and field experience. The paper concludes that computerized procedures are ready for pilot applications at stations. (author)

  1. Interactive topology optimization on hand-held devices

    DEFF Research Database (Denmark)

    Aage, Niels; Nobel-Jørgensen, Morten; Andreasen, Casper Schousboe

    2013-01-01

    This paper presents an interactive topology optimization application designed for hand-held devices running iOS or Android. The TopOpt app solves the 2D minimum compliance problem with interactive control of load and support positions as well as volume fraction. Thus, it is possible to change......OS devices from the Apple App Store, at Google Play for the Android platform, and a web-version can be run from www.topopt.dtu.dk....

  2. X-ray astronomy

    International Nuclear Information System (INIS)

    Giacconi, R.; Gursky, H.

    1974-01-01

    This text contains ten chapters and three appendices. Following an introduction, chapters two through five deal with observational techniques, mechanisms for the production of x rays in a cosmic setting, the x-ray sky and solar x-ray emission. Chapters six through ten include compact x-ray sources, supernova remnants, the interstellar medium, extragalactic x-ray sources and the cosmic x-ray background. Interactions of x rays with matter, units and conversion factors and a catalog of x-ray sources comprise the three appendices. (U.S.)

  3. X-ray equipment for dental diagnostics with a dose-rate controller influencing the exposure voltage

    International Nuclear Information System (INIS)

    Franke, K.

    1980-01-01

    If a synoptic picture is taken of teeth or jaw the X-ray tube and the film carrier with the film and a radiation detector are moving around the patient's head. The detector signal is connected to the input of the actual value of the dose-rate controller containing a scanning and stop circuit with a storing time roughly corresponding to the exposure of three teeth. After the shot the actual value of the dose rate is adjusted to the desired value, thus a constant quality of the following pictures being achieved. (RW) [de

  4. Feasibility study for application of the compressed-sensing framework to interior computed tomography (ICT) for low-dose, high-accurate dental x-ray imaging

    Science.gov (United States)

    Je, U. K.; Cho, H. M.; Cho, H. S.; Park, Y. O.; Park, C. K.; Lim, H. W.; Kim, K. S.; Kim, G. A.; Park, S. Y.; Woo, T. H.; Choi, S. I.

    2016-02-01

    In this paper, we propose a new/next-generation type of CT examinations, the so-called Interior Computed Tomography (ICT), which may presumably lead to dose reduction to the patient outside the target region-of-interest (ROI), in dental x-ray imaging. Here an x-ray beam from each projection position covers only a relatively small ROI containing a target of diagnosis from the examined structure, leading to imaging benefits such as decreasing scatters and system cost as well as reducing imaging dose. We considered the compressed-sensing (CS) framework, rather than common filtered-backprojection (FBP)-based algorithms, for more accurate ICT reconstruction. We implemented a CS-based ICT algorithm and performed a systematic simulation to investigate the imaging characteristics. Simulation conditions of two ROI ratios of 0.28 and 0.14 between the target and the whole phantom sizes and four projection numbers of 360, 180, 90, and 45 were tested. We successfully reconstructed ICT images of substantially high image quality by using the CS framework even with few-view projection data, still preserving sharp edges in the images.

  5. Crystallite size and lattice distortion of human dental enamel estimated from the integral width of x-ray diffraction peak profile

    International Nuclear Information System (INIS)

    Maruyama, Fumiaki; Sakae, Toshiro

    2000-01-01

    Crystallite size and lattice distortion of human dental enamel were estimated by peak profile analysis using x-ray diffraction pattern. Firstly, noises were removed from x-ray diffraction pattern, and deconvolution of overlapping peaks and determination of baseline level were carried out. Then, the instrumental peak broadening and effect of overlapping Kα1 and Kα2 were eliminated to obtain pure peak profile using the Stokes's Fourier method. The integral width method was applied for estimation of crystallite size and 'upper-limit of distortion', assuming the peak profile as Cauchy function. The estimated crystallite size and distortion were ca. 210 A and ca. 0.4% in the a-axis direction and ca. 550 A and ca. 0.7% in the c-axis direction, respectively. The crystallite size value along the a-axis was almost the same to the previously reported values, but the value for along the c-axis was nearly half of the reported values. The crystallite size in this study means the size of coherent domain in contrast to the size of particle which may contain several domains in the case of enamel crystals. The results suggest that human enamel crystals grow in their size along the c-axis by multiplication, fusion of crystallites. It was notable that the distortion value was larger in the c-axis direction. The phenomenon may partly be due to the high carbonate ion content of enamel crystals and partly due to crystal growth mechanism. (author)

  6. Chest X-Ray

    Medline Plus

    Full Text Available ... about chest radiography also known as chest x-rays. Chest x-rays are the most commonly performed x-ray exams and use a very small dose of ... of the inside of the chest. A chest x-ray is used to evaluate the lungs, heart and ...

  7. Radiation protection in dentistry. Recommended safety procedures for the use of dental x-ray equipment. Safety code 30

    International Nuclear Information System (INIS)

    1994-01-01

    The Radiation Protection Bureau has prepared a series of documents on safety codes to set out requirements for the safe use of radiation-emitting equipment. This Safety Code has been prepared to provide specific guidance to the dentist, dental hygienist, dental assistant and other support personnel concerned with safety procedures and equipment performance. Dental radiography is one of the most valuable tools used in modern dental health care. It makes possible the diagnosis of physical conditions that would otherwise be difficult to identify. The use of dental radiological procedures must be carefully managed, because x-radiation has the potential for damaging healthy cells and tissues. Although no known occurrence of cancer or genetic damage has been observed from radiation doses delivered in modern dentistry, and until more evidence is available, one should practice radiation hygiene with the same care as would be dictated if a hazard were known to exist. The aim of radiation protection in dentistry is to obtain the desired clinical information with minimal radiation exposure to patients, dental personnel and the public. 15 tabs

  8. Radiation protection in dentistry. Recommended safety procedures for the use of dental x-ray equipment. Safety code 30

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1994-12-31

    The Radiation Protection Bureau has prepared a series of documents on safety codes to set out requirements for the safe use of radiation-emitting equipment. This Safety Code has been prepared to provide specific guidance to the dentist, dental hygienist, dental assistant and other support personnel concerned with safety procedures and equipment performance. Dental radiography is one of the most valuable tools used in modern dental health care. It makes possible the diagnosis of physical conditions that would otherwise be difficult to identify. The use of dental radiological procedures must be carefully managed, because x-radiation has the potential for damaging healthy cells and tissues. Although no known occurrence of cancer or genetic damage has been observed from radiation doses delivered in modern dentistry, and until more evidence is available, one should practice radiation hygiene with the same care as would be dictated if a hazard were known to exist. The aim of radiation protection in dentistry is to obtain the desired clinical information with minimal radiation exposure to patients, dental personnel and the public. 15 tabs.

  9. X-ray sky

    International Nuclear Information System (INIS)

    Gruen, M.; Koubsky, P.

    1977-01-01

    The history is described of the discoveries of X-ray sources in the sky. The individual X-ray detectors are described in more detail, i.e., gas counters, scintillation detectors, semiconductor detectors, and the principles of X-ray spectrometry and of radiation collimation aimed at increased resolution are discussed. Currently, over 200 celestial X-ray sources are known. Some were identified as nebulae, in some pulsations were found or the source was identified as a binary star. X-ray bursts of novae were also observed. The X-ray radiation is briefly mentioned of spherical star clusters and of extragalactic X-ray sources. (Oy)

  10. A hand-held robotic device for peripheral intravenous catheterization.

    Science.gov (United States)

    Cheng, Zhuoqi; Davies, Brian L; Caldwell, Darwin G; Barresi, Giacinto; Xu, Qinqi; Mattos, Leonardo S

    2017-12-01

    Intravenous catheterization is frequently required for numerous medical treatments. However, this process is characterized by a high failure rate, especially when performed on difficult patients such as newborns and infants. Very young patients have small veins, and that increases the chances of accidentally puncturing the catheterization needle directly through them. In this article, we present the design, development and experimental evaluation of a novel hand-held robotic device for improving the process of peripheral intravenous catheterization by facilitating the needle insertion procedure. To our knowledge, this design is the first hand-held robotic device for assisting in the catheterization insertion task. Compared to the other available technologies, it has several unique advantages such as being compact, low-cost and able to reliably detect venipuncture. The system is equipped with an electrical impedance sensor at the tip of the catheterization needle, which provides real-time measurements used to supervise and control the catheter insertion process. This allows the robotic system to precisely position the needle within the lumen of the target vein, leading to enhanced catheterization success rate. Experiments conducted to evaluate the device demonstrated that it is also effective to deskill the task. Naïve subjects achieved an average catheterization success rate of 88% on a 1.5 mm phantom vessel with the robotic device versus 12% with the traditional unassisted system. The results of this work prove the feasibility of a hand-held assistive robotic device for intravenous catheterization and show that such device has the potential to greatly improve the success rate of these difficult operations.

  11. Adaptive RF front-ends for hand-held applications

    CERN Document Server

    van Bezooijen, Andre; van Roermund, Arthur

    2010-01-01

    The RF front-end - antenna combination is a vital part of a mobile phone because its performance is very relevant to the link quality between hand-set and cellular network base-stations. The RF front-end performance suffers from changes in operating environment, like hand-effects, that are often unpredictable. ""Adaptive RF Front-Ends for Hand-Held Applications"" presents an analysis on the impact of fluctuating environmental parameters. In order to overcome undesired behavior two different adaptive control methods are treated that make RF frond-ends more resilient: adaptive impedance control,

  12. Development of dual sensor hand-held detector

    Science.gov (United States)

    Sezgin, Mehmet

    2010-04-01

    In this paper hand-held dual sensor detector development requirements are considered dedicated to buried object detection. Design characteristics of such a system are categorized and listed. Hardware and software structures, ergonomics, user interface, environmental and EMC/EMI tests to be applied and performance test issues are studied. Main properties of the developed system (SEZER) are presented, which contains Metal Detector (MD) and Ground Penetrating Radar (GPR). The realized system has ergonomic structure and can detect both metallic and non-metallic buried objects. Moreover classification of target is possible if it was defined to the signal processing software in learning phase.

  13. Why Do I Need X-Rays?

    Science.gov (United States)

    ... Baby Teeth? What is a Composite Resin (White Filling)? What is Orofacial Pain? Learn what those dental ... jaw placement and the overall composition of your facial bones. X-rays can help your dentist determine ...

  14. A hand-held beta imaging probe for FDG.

    Science.gov (United States)

    Singh, Bipin; Stack, Brendan C; Thacker, Samta; Gaysinskiy, Valeriy; Bartel, Twyla; Lowe, Val; Cool, Steven; Entine, Gerald; Nagarkar, Vivek

    2013-04-01

    Advances in radiopharmaceuticals and clinical understanding have escalated the use of intraoperative gamma probes in surgery. However, most probes on the market are non-imaging gamma probes that suffer from the lack of ancillary information of the surveyed tissue area. We have developed a novel, hand-held digital Imaging Beta Probe™ (IBP™) to be used in surgery in conjunction with beta-emitting radiopharmaceuticals such as (18)FDG, (131)I and (32)P for real-time imaging of a surveyed area with higher spatial resolution and sensitivity and greater convenience than existing instruments. We describe the design and validation of a hand-held beta probe intended to be used as a visual mapping device to locate and confirm excision of (18)FDG-avid primary tumors and metastases in an animal model. We have demonstrated a device which can generate beta images from (18)FDG avid lesions in an animal model. It is feasible to image beta irradiation in animal models of cancer given (18)FDG. This technology may be applied to clinical mapping of tumors and/or their metastases in the operating room. Visual image depiction of malignancy may aid the surgeon in localization and excision of lesions of interest.

  15. Measuring thyroid uptake with hand-held radiation monitors

    International Nuclear Information System (INIS)

    Deschamps, M.

    1987-04-01

    With the use of Iodine 123, 125 and 131 and some compounds of Technetium-99 m, a fraction of the isotopes can be trapped in the thyroid of the technicians. We used the hand-held radiation contamination or survey meters of the nine (9) Nuclear medicine departments we visited to see if they were adequate for the evaluation of thyroid uptake of the users. Measurements on a neck-phanton helped us to determine a minimum detectable activity for each isotope. We were then able to check if the measurements of investigations and action levels were possible. None of the hand-held radiation monitors are completely satisfactory for the measure of thyroid uptake of the user. We discuss a class of equipment capable of measuring radiation emissions at the investigation level. Measurement at the action level is possible with meters having scintillation or proportional probes but none of them permits the discrimination in energy required for a quantitative evaluation of the radioisotopes used

  16. Hand held data collection and monitoring system for nuclear facilities

    Science.gov (United States)

    Brayton, D.D.; Scharold, P.G.; Thornton, M.W.; Marquez, D.L.

    1999-01-26

    Apparatus and method is disclosed for a data collection and monitoring system that utilizes a pen based hand held computer unit which has contained therein interaction software that allows the user to review maintenance procedures, collect data, compare data with historical trends and safety limits, and input new information at various collection sites. The system has a means to allow automatic transfer of the collected data to a main computer data base for further review, reporting, and distribution purposes and uploading updated collection and maintenance procedures. The hand held computer has a running to-do list so sample collection and other general tasks, such as housekeeping are automatically scheduled for timely completion. A done list helps users to keep track of all completed tasks. The built-in check list assures that work process will meet the applicable processes and procedures. Users can hand write comments or drawings with an electronic pen that allows the users to directly interface information on the screen. 15 figs.

  17. Hand held data collection and monitoring system for nuclear facilities

    International Nuclear Information System (INIS)

    Brayton, D.D.; Scharold, P.G.; Thornton, M.W.; Marquez, D.L.

    1999-01-01

    Apparatus and method is disclosed for a data collection and monitoring system that utilizes a pen based hand held computer unit which has contained therein interaction software that allows the user to review maintenance procedures, collect data, compare data with historical trends and safety limits, and input new information at various collection sites. The system has a means to allow automatic transfer of the collected data to a main computer data base for further review, reporting, and distribution purposes and uploading updated collection and maintenance procedures. The hand held computer has a running to-do list so sample collection and other general tasks, such as housekeeping are automatically scheduled for timely completion. A done list helps users to keep track of all completed tasks. The built-in check list assures that work process will meet the applicable processes and procedures. Users can hand write comments or drawings with an electronic pen that allows the users to directly interface information on the screen. 15 figs

  18. Intraoral radiology in general dental practices. A comparison of digital and film-based X-ray systems with regard to radiation protection and dose reduction

    Energy Technology Data Exchange (ETDEWEB)

    Anissi, H.D. [Ulm Univ. (Germany). Dentistry; Geibel, M.A. [Ulm Univ. (Germany). Dept. of Dentomaxillofacial Surgery

    2014-08-15

    Purpose: The purpose of this study was to gain insight into the distribution and application of digital intraoral radiographic techniques within general dental practices and to compare these with film-based systems in terms of patient dose reduction. Materials and Methods: 1100 questionnaires were handed out to general dental practitioners. Data was analyzed with respect to the type of system by using descriptive statistics and nonparametric tests, i.e. Kruskal-Wallis, Mann-Whitney and chi-square test (SPSS 20). Results: 64% of the questioned dentists still use film-based radiology, 23% utilize storage phosphor plate (SPP) systems and 13% use a charge-coupled device (CCD). A strong correlation between the number of dentists working in a practice and the use of digital dental imaging was observed. Almost 3/4 of the film users work with E- or F-speed film. 45% of them refuse to change to a digital system. The use of lead aprons was popular, while only a minority preferred thyroid shields and rectangular collimators. A fourfold reduction of exposure time from D-speed film to CCD systems was observed. Due to detector size and positioning errors, users of CCD systems take significantly more single-tooth radiographs in total. Considering the number of radiographs per patient, there is only a slight tendency towards more X-rays with CCD systems. Up to image generation, digital systems seem to be as or even more difficult to handle than film-based systems, while their handling was favored after radiographic exposure. Conclusion: Despite a slight increase of radiographs taken with CCD systems, there is a significant dosage reduction. Corresponding to the decrease in exposure time, the patient dose for SPP systems is reduced to one half compared to film. The main issues in CCD technology are positioning errors and the size of the X-ray detectors which are difficult to eliminate. The usage of radiation protection measures still needs to be improved. (orig.)

  19. Intraoral radiology in general dental practices - a comparison of digital and film-based X-ray systems with regard to radiation protection and dose reduction.

    Science.gov (United States)

    Anissi, H D; Geibel, M A

    2014-08-01

    The purpose of this study was to gain insight into the distribution and application of digital intraoral radiographic techniques within general dental practices and to compare these with film-based systems in terms of patient dose reduction. 1100 questionnaires were handed out to general dental practitioners. Data was analyzed with respect to the type of system by using descriptive statistics and nonparametric tests, i.e. Kruskal-Wallis, Mann-Whitney and chi-square test (SPSS 20). 64% of the questioned dentists still use film-based radiology, 23% utilize storage phosphor plate (SPP) systems and 13% use a charge-coupled device (CCD). A strong correlation between the number of dentists working in a practice and the use of digital dental imaging was observed. Almost 3/4 of the film users work with E- or F-speed film. 45% of them refuse to change to a digital system. The use of lead aprons was popular, while only a minority preferred thyroid shields and rectangular collimators. A fourfold reduction of exposure time from D-speed film to CCD systems was observed. Due to detector size and positioning errors, users of CCD systems take significantly more single-tooth radiographs in total. Considering the number of radiographs per patient, there is only a slight tendency towards more X-rays with CCD systems. Up to image generation, digital systems seem to be as or even more difficult to handle than film-based systems, while their handling was favored after radiographic exposure. Despite a slight increase of radiographs taken with CCD systems, there is a significant dosage reduction. Corresponding to the decrease in exposure time, the patient dose for SPP systems is reduced to one half compared to film. The main issues in CCD technology are positioning errors and the size of the X-ray detectors which are difficult to eliminate. The usage of radiation protection measures still needs to be improved. ► Responsible use of digital intraoral radiology results in a significant

  20. Intraoral radiology in general dental practices. A comparison of digital and film-based X-ray systems with regard to radiation protection and dose reduction

    International Nuclear Information System (INIS)

    Anissi, H.D.; Geibel, M.A.

    2014-01-01

    Purpose: The purpose of this study was to gain insight into the distribution and application of digital intraoral radiographic techniques within general dental practices and to compare these with film-based systems in terms of patient dose reduction. Materials and Methods: 1100 questionnaires were handed out to general dental practitioners. Data was analyzed with respect to the type of system by using descriptive statistics and nonparametric tests, i.e. Kruskal-Wallis, Mann-Whitney and chi-square test (SPSS 20). Results: 64% of the questioned dentists still use film-based radiology, 23% utilize storage phosphor plate (SPP) systems and 13% use a charge-coupled device (CCD). A strong correlation between the number of dentists working in a practice and the use of digital dental imaging was observed. Almost 3/4 of the film users work with E- or F-speed film. 45% of them refuse to change to a digital system. The use of lead aprons was popular, while only a minority preferred thyroid shields and rectangular collimators. A fourfold reduction of exposure time from D-speed film to CCD systems was observed. Due to detector size and positioning errors, users of CCD systems take significantly more single-tooth radiographs in total. Considering the number of radiographs per patient, there is only a slight tendency towards more X-rays with CCD systems. Up to image generation, digital systems seem to be as or even more difficult to handle than film-based systems, while their handling was favored after radiographic exposure. Conclusion: Despite a slight increase of radiographs taken with CCD systems, there is a significant dosage reduction. Corresponding to the decrease in exposure time, the patient dose for SPP systems is reduced to one half compared to film. The main issues in CCD technology are positioning errors and the size of the X-ray detectors which are difficult to eliminate. The usage of radiation protection measures still needs to be improved. (orig.)

  1. Chest X-Ray

    Medline Plus

    Full Text Available ... some concerns about chest x-rays. However, it’s important to consider the likelihood of benefit to your health. While a chest x-ray use a ... posted: How to Obtain and Share ...

  2. Chest X-Ray

    Medline Plus

    Full Text Available ... X-ray Transcript Welcome to Radiology Info dot org! Hello, I’m Dr. Geoffrey Rubin, a radiologist ... about chest x-rays, visit Radiology Info dot org. Thank you for your time! Spotlight Recently posted: ...

  3. X-ray apparatus

    International Nuclear Information System (INIS)

    Sell, L.J.

    1981-01-01

    A diagnostic x-ray device, readily convertible between conventional radiographic and tomographic operating modes, is described. An improved drive system interconnects and drives the x-ray source and the imaging device through coordinated movements for tomography

  4. X-ray - skeleton

    Science.gov (United States)

    ... this page: //medlineplus.gov/ency/article/003381.htm X-ray - skeleton To use the sharing features on this ... Degenerative bone conditions Osteomyelitis Risks There is low radiation exposure. X-rays machines are set to provide the smallest ...

  5. Hand-held spectrophotometer design for textile fabrics

    Science.gov (United States)

    Böcekçi, Veysel Gökhan; Yıldız, Kazım

    2017-09-01

    In this study, a hand-held spectrophotometer was designed by taking advantage of the developments in modern optoelectronic technology. Spectrophotometer devices are used to determine the color information from the optic properties of the materials. As an alternative to a desktop spectrophotometer device we have implemented, it is the first prototype, low cost and portable. The prototype model designed for the textile industry can detect the color tone of any fabric. The prototype model consists of optic sensor, processor, display floors. According to the color applied on the optic sensor, it produces special frequency information on its output at that color value. In Arduino type processor, the frequency information is evaluated by the program we have written and the color tone information between 0-255 ton is decided and displayed on the screen.

  6. Comparative Geometrical Investigations of Hand-Held Scanning Systems

    Science.gov (United States)

    Kersten, T. P.; Przybilla, H.-J.; Lindstaedt, M.; Tschirschwitz, F.; Misgaiski-Hass, M.

    2016-06-01

    An increasing number of hand-held scanning systems by different manufacturers are becoming available on the market. However, their geometrical performance is little-known to many users. Therefore the Laboratory for Photogrammetry & Laser Scanning of the HafenCity University Hamburg has carried out geometrical accuracy tests with the following systems in co-operation with the Bochum University of Applied Sciences (Laboratory for Photogrammetry) as well as the Humboldt University in Berlin (Institute for Computer Science): DOTProduct DPI-7, Artec Spider, Mantis Vision F5 SR, Kinect v1 + v2, Structure Sensor and Google's Project Tango. In the framework of these comparative investigations geometrically stable reference bodies were used. The appropriate reference data were acquired by measurement with two structured light projection systems (AICON smartSCAN and GOM ATOS I 2M). The comprehensive test results of the different test scenarios are presented and critically discussed in this contribution.

  7. COMPARATIVE GEOMETRICAL INVESTIGATIONS OF HAND-HELD SCANNING SYSTEMS

    Directory of Open Access Journals (Sweden)

    T. P. Kersten

    2016-06-01

    Full Text Available An increasing number of hand-held scanning systems by different manufacturers are becoming available on the market. However, their geometrical performance is little-known to many users. Therefore the Laboratory for Photogrammetry & Laser Scanning of the HafenCity University Hamburg has carried out geometrical accuracy tests with the following systems in co-operation with the Bochum University of Applied Sciences (Laboratory for Photogrammetry as well as the Humboldt University in Berlin (Institute for Computer Science: DOTProduct DPI-7, Artec Spider, Mantis Vision F5 SR, Kinect v1 + v2, Structure Sensor and Google’s Project Tango. In the framework of these comparative investigations geometrically stable reference bodies were used. The appropriate reference data were acquired by measurement with two structured light projection systems (AICON smartSCAN and GOM ATOS I 2M. The comprehensive test results of the different test scenarios are presented and critically discussed in this contribution.

  8. Chest X-Ray

    Medline Plus

    Full Text Available ... I’d like to talk with you about chest radiography also known as chest x-rays. Chest x-rays are the most ... far outweighs any risk. For more information about chest x-rays, visit Radiology Info dot org. Thank you for your time! ...

  9. Chest X-Ray

    Medline Plus

    Full Text Available ... by Image/Video Gallery Your Radiologist Explains Chest X-ray Transcript Welcome to Radiology Info dot org! Hello, ... you about chest radiography also known as chest x-rays. Chest x-rays are the most commonly performed ...

  10. Flash X-ray

    International Nuclear Information System (INIS)

    Sato, Eiichi

    2003-01-01

    Generation of quasi-monochromatic X-ray by production of weakly ionized line plasma (flash X-ray), high-speed imaging by the X-ray and high-contrast imaging by the characteristic X-ray absorption are described. The equipment for the X-ray is consisted from the high-voltage power supply and condenser, turbo molecular pump, and plasma X-ray tube. The tube has a long linear anticathode to produce the line plasma and flash X-ray at 20 kA current at maximum. X-ray spectrum is measured by the imaging plate equipped in the computed radiography system after diffracted by a LiF single crystal bender. Cu anticathode generates sharp peaks of K X-ray series. The tissue images are presented for vertebra, rabbit ear and heart, and dog heart by X-ray fluoroscopy with Ce anticathode. Generation of K-orbit characteristic X-ray with extremely low bremsstrahung is to be attempted for medical use. (N.I.)

  11. Adaptation Measurement of CAD/CAM Dental Crowns with X-Ray Micro-CT: Metrological Chain Standardization and 3D Gap Size Distribution

    Directory of Open Access Journals (Sweden)

    L. Tapie

    2016-01-01

    Full Text Available Computer-Aided Design and Manufacturing systems are increasingly used to produce dental prostheses, but the parts produced suffer from a lack of evaluation, especially concerning the internal gap of the final assembly, that is, the space between the prepared tooth and the prosthesis. X-ray micro-Computed Tomography (micro-CT is a noninvasive imaging technique enabling the internal inspection of the assembly. It has proved to be an efficient tool for measuring the gap. In this study, a critical review of the protocols using micro-CT to quantify the gap is proposed as an introduction to a new protocol aimed at minimizing errors and enabling comparison between CAD/CAM systems. To compare different systems, a standardized protocol is proposed including two reference geometries. Micro-CT is used to acquire the reference geometries. A new 3D method is then proposed and a new indicator is defined (Gap Size Distribution (GSD. In addition, the usual 2D measurements are described and discussed. The 3D gap measurement method proposed can be used in clinical case geometries and has the considerable advantage of minimizing the data processing steps before performing the measurements.

  12. High energy X-ray diffraction study of a dental ceramics–titanium functional gradient material prepared by field assisted sintering technique

    International Nuclear Information System (INIS)

    Witte, K.; Bodnar, W.; Schell, N.; Lang, H.; Burkel, E.

    2014-01-01

    A functional gradient material with eleven layers composed of a dental ceramics and titanium was successfully consolidated using field assisted sintering technique in a two-step sintering process. High energy X-ray diffraction studies on the gradient were performed at High Energy Material Science beamline at Desy in Hamburg. Phase composition, crystal unit edges and lattice mismatch along the gradient were determined applying Rietveld refinement procedure. Phase analysis revealed that the main crystalline phase present in the gradient is α-Ti. Crystallinity increases stepwisely along the gradient with a decreasing increment between every next layer, following rather the weight fraction of titanium. The crystal unit edge a of titanium remains approximately constant with a value of 2.9686(1) Å, while c is reduced with increasing amount of titanium. In the layer with pure titanium the crystal unit edge c is constant with a value of 4.7174(2) Å. The lattice mismatch leading to an internal stress was calculated over the whole gradient. It was found that the maximal internal stress in titanium embedded in the studied gradient is significantly smaller than its yield strength, which implies that the structure of titanium along the whole gradient is mechanically stable. - Highlights: • High energy XRD studies of dental ceramics–Ti gradient material consolidated by FAST. • Phase composition, crystallinity and lattice parameters are determined. • Crystallinity increases stepwisely along the gradient following weight fraction of Ti. • Lattice mismatch leading to internal stress is calculated over the whole gradient. • Internal stress in α-Ti embedded in the gradient is smaller than its yield strength

  13. Development of the three dimensional image display program for limited cone beam X-ray CT for dental use (Ortho-CT)

    International Nuclear Information System (INIS)

    Arai, Yoshinori; Hashimoto, Koji; Shinoda, Koji

    2000-01-01

    We have already developed and reported a limited cone beam X-ray CT system for dental use (Ortho-CT). This system has been used clinically since 1997. In this study, we report a 3D surface display program for Ortho-CT which has been newly-developed by the authors. The 3D surface display software has been developed using visual C ++ (Microsoft Co. WA. USA) and a personal computer (Pentium 450MHz Intel Co. CA USA, Windows NT 4.0 Microsoft WA. USA). In this software, the 3D surface images are recorded as AVI files and can be displayed on the personal computer. The 3D images can be rotated and a stepwise change of the threshold voxel value for binary image formation can be automatically used. We have applied these 3D surface images to clinical studies from January 1999 to May 1999 at the Radiology section in our Dental hospital. The images can be displayed very easily in personal computers using AVI files. Thirty-five cases have been reconstructed using 3D surface images in this way. The 3D surface image is useful in the diagnosis of fractures of the mandibular head and impacted teeth. Only teeth are observed when a relative threshold voxel value is set at a high level such as about 0.37. When the threshold is changed to a lower value (about 0.3), we can observe both teeth and the surface of the bone. We have developed a 3D surface display program for personal computers. The images are useful for the diagnosis of the pathosis in the maxillofacial region. (author)

  14. Identification of pulpitis at dental X-ray periapical radiography based on edge detection, texture description and artificial neural networks

    Directory of Open Access Journals (Sweden)

    Bernard Y Tumbelaka

    2014-01-01

    Full Text Available Objectives: The aim of the present research was to identify pulpitis through periapical radiography by applying edges as basis image features, the texture description and the artificial neural networks (ANNs. Materials and Methods: Input image data records of 10 molar and 10 canine teeth were used. The clinical diagnosis of interest cases were represented as normal pulp, reversible and irreversible pulpitis, and necrotic pulp. The following image processing steps were done. First, the data records were converted digitally and preprocessed as its original image using the Gaussian Filter to obtain the best smoothed intensity distribution. Second, the local image differentiation was used to produce edge detector operators, e(x,y as the image gradient; ∇f(x,y providing useful information about the local intensity variations. Third, these results were analyzed by using the texture descriptors to obtain digitally the image entropy, H. The fourth step, all were characterized by the ANNs. Results: The edge detection carried important information about the object boundaries of pulpal health and pain conditions in the dental pulp significantly. The image entropy which was identified, the diagnostic term, was obtained from texture descriptors in the segmentation regions where the curves of pulp states tent convergence with the normal pulp line from 4.9014 to 4.6843 decreasing to the reversible and the irreversible pulpitis line include the nectrotic pulp line from 4.6812 to 4.5926 and then inputting to the ANNs analysis at the same of mean square error around 0.0003. Conclusions: Referred to these results, the correlation of the image entropy and the ANNs analysis could be linearly classified with the critical point of 4.6827. Finally, it could be concluded that the direct reading radiography is better to be digitized in order to provide us the best choice for diagnosis validation.

  15. X-ray astronomy

    International Nuclear Information System (INIS)

    Culhane, J.L.; Sanford, P.W.

    1981-01-01

    X-ray astronomy has been established as a powerful means of observing matter in its most extreme form. The energy liberated by sources discovered in our Galaxy has confirmed that collapsed stars of great density, and with intense gravitational fields, can be studied by making observations in the X-ray part of the electromagnetic spectrum. The astronomical objects which emit detectable X-rays include our own Sun and extend to quasars at the edge of the Universe. This book describes the history, techniques and results obtained in the first twenty-five years of exploration. Space rockets and satellites are essential for carrying the instruments above the Earth's atmosphere where it becomes possible to view the X-rays from stars and nebulae. The subject is covered in chapters, entitled: the birth of X-ray astronomy; the nature of X-radiation; X-rays from the Sun; solar-flare X-rays; X-rays from beyond the solar system; supernovae and their remnants; X-rays from binary stars; white dwarfs and neutron stars; black holes; X-rays from galaxies and quasars; clusters of galaxies; the observatories of the future. (author)

  16. Diagnostic x-ray equipment and its use

    International Nuclear Information System (INIS)

    1993-01-01

    The guide contains the requirements for the use, structure and operation of diagnostic x-ray equipment necessary to ensure the radiation safety of workers and patients. The guide is intended for the party responsible for operating the x-ray units, the staff operating, installing and maintaining x-ray examination equipment, and companies manufacturing and selling such equipment. Special requirements for mammographic and dental x-ray equipments are given in separate ST guides. (5 refs., 1 figs., 2 tabs.)

  17. Attentionally splitting the mass distribution of hand-held rods.

    Science.gov (United States)

    Burton, G; Turvey, M T

    1991-08-01

    Two experiments on the length-perception capabilities of effortful or dynamic touch differed only in terms of what the subject intended to perceive, while experimental conditions and apparatus were held constant. In each trial, a visually occluded rod was held as still as possible by the subject at an intermediate position. For two thirds of the trials, a weight was attached to the rod above or below the hand. In Experiment 1, in which the subject's task was to perceive the distance reachable with the portion of the rod forward of the hand, perceived extent was a function of the first moment of the mass distribution associated with the forward portion of the rod, and indifferent to the first moment of the entire rod. In Experiment 2, in which the task was to perceive the distance reachable with the entire rod if it was held at an end, the pattern of results was reversed. These results indicate the capability of selective sensitivity to different aspects of a hand-held object's mass distribution, without the possibility of differential exploration specific to these two tasks. Results are discussed in relation to possible roles of differential information, intention, and self-organization in the explanations of selective perceptual abilities.

  18. Guides for intraoral x-rays

    International Nuclear Information System (INIS)

    Ogunsunlade, O.A.

    1988-01-01

    An h-shaped exterior guide for use in combination with a SNAP-A-RAY film holder for accurately aligning a beam from an X-ray cone with an X-ray film during the process of taking intraoral periapical dental X-rays of the maxillary and mandibular teeth is described comprising: a first guide arm laterally and detachably connectable through a housing means; a traverse arm extending from the midpoint of the first guide arm and parallel to the X-ray film; and a second guide arm extending perpendicularly from an end of the traverse arm toward a plane of the X-ray film and in parallel relation up to an end point of the first guide arm

  19. X-ray holography

    International Nuclear Information System (INIS)

    Faigel, G.; Tegze, M.; Belakhovsky, M.; Marchesini, S.; Bortel, G.

    2003-01-01

    In the last decade holographic methods using hard X-rays were developed. They are able to resolve atomic distances, and can give the 3D arrangement of atoms around a selected element. Therefore, hard X-ray holography has potential applications in chemistry, biology and physics. In this article we give a general description of these methods and discuss the developments in the experimental technique. The capabilities of hard X-ray holography are demonstrated by examples

  20. Providing x-rays

    International Nuclear Information System (INIS)

    Mallozzi, P.J.; Epstein, H.M.

    1985-01-01

    This invention provides an apparatus for providing x-rays to an object that may be in an ordinary environment such as air at approximately atmospheric pressure. The apparatus comprises: means (typically a laser beam) for directing energy onto a target to produce x-rays of a selected spectrum and intensity at the target; a fluid-tight enclosure around the target; means for maintaining the pressure in the first enclosure substantially below atmospheric pressure; a fluid-tight second enclosure adjoining the first enclosure, the common wall portion having an opening large enough to permit x-rays to pass through but small enough to allow the pressure reducing means to evacuate gas from the first enclosure at least as fast as it enters through the opening; the second enclosure filled with a gas that is highly transparent to x-rays; the wall of the second enclosure to which the x-rays travel having a portion that is highly transparent to x-rays (usually a beryllium or plastic foil), so that the object to which the x-rays are to be provided may be located outside the second enclosure and adjacent thereto and thus receive the x-rays substantially unimpeded by air or other intervening matter. The apparatus is particularly suited to obtaining EXAFS (extended x-ray fine structure spectroscopy) data on a material

  1. New intraoral x-ray fluorographic imaging for dentistry

    International Nuclear Information System (INIS)

    Higashi, T.; Osada, T.; Aoyama, W.; Iguchi, M.; Suzuki, S.; Kanno, M.; Moriya, K.; Yoshimura, M.; Tusuda, M.

    1983-01-01

    A new dental x-ray fluorographic unit has been developed. This unit is composed of small intraoral x-ray tube, a compact x-ray image intensifier, and a high-resolution TV system. The purposes for developing this equipment were to (1) directly observe the tooth during endodontic procedures and (2) reduce x-ray exposure to the patient and the dentist. The radiation exposure can be reduced to about 1/600 the exposure used with conventional dental film. In clinical trials, a satisfactory fluorographic dental image for endodontic treatment was obtained with this new device

  2. Portable Hand-Held Electrochemical Sensor for the Transuranics

    Energy Technology Data Exchange (ETDEWEB)

    Dale D. Russell, William B. Knowlton, Ph.D.; Russel Hertzog, Ph.D

    2005-11-25

    sensitive detector for uranium. Millimeter scale electrodes, operated by a hand-held instrument assembled in this lab and operated in the voltammetric mode, were transported to the DOE-Nevada test site (Las Vegas, NV) where field detection and quantitation of plutonium, uranium, and a mixture of these two elements was also demonstrated. Several probe designs were prepared, built and tested including probes with movable protective windows. A miniature, battery powered potentiostat was designed, built and demonstrated for use in a hand-held field portable instrument. This work was performed largely by undergraduates who gained valuable research experience, and many of them have continued on to graduate schools. In addition, they all gained exposure to and appreciation for national security research, in particular non-proliferation research. Four graduate students participated and one earned the MS degree on this project.

  3. Direction-Sensitive Hand-Held Gamma-Ray Spectrometer

    Energy Technology Data Exchange (ETDEWEB)

    Mukhopadhyay, S.

    2012-10-04

    A novel, light-weight, hand-held gamma-ray detector with directional sensitivity is being designed. The detector uses a set of multiple rings around two cylindrical surfaces, which provides precise location of two interaction points on two concentric cylindrical planes, wherefrom the source location can be traced back by back projection and/or Compton imaging technique. The detectors are 2.0 × 2.0 mm europium-doped strontium iodide (SrI2:Eu2+) crystals, whose light output has been measured to exceed 120,000 photons/MeV, making it one of the brightest scintillators in existence. The crystal’s energy resolution, less than 3% at 662 keV, is also excellent, and the response is highly linear over a wide range of gamma-ray energies. The emission of SrI2:Eu2+ is well matched to both photo-multiplier tubes and blue-enhanced silicon photodiodes. The solid-state photomultipliers used in this design (each 2.0 × 2.0 mm) are arrays of active pixel sensors (avalanche photodiodes driven beyond their breakdown voltage in reverse bias); each pixel acts as a binary photon detector, and their summed output is an analog representation of the total photon energy, while the individual pixel accurately defines the point of interaction. A simple back-projection algorithm involving cone-surface mapping is being modeled. The back projection for an event cone is a conical surface defining the possible location of the source. The cone axis is the straight line passing through the first and second interaction points.

  4. X-ray interferometers

    International Nuclear Information System (INIS)

    Franks, A.

    1980-01-01

    An improved type of amplitude-division x-ray interferometer is described. The wavelength at which the interferometer can operate is variable, allowing the instrument to be used to measure x-ray wavelength, and the angle of inclination is variable for sample investigation. (U.K.)

  5. Extremity x-ray

    Science.gov (United States)

    ... page: //medlineplus.gov/ency/article/003461.htm Extremity x-ray To use the sharing features on this page, ... in the body Risks There is low-level radiation exposure. X-rays are monitored and regulated to provide the ...

  6. X-rays utilization

    International Nuclear Information System (INIS)

    Rebigan, F.

    1979-03-01

    The modality of X-ray utilization in different activities and economy is given. One presents firstly quantities and units used in radiation dosimetry and other fields. One gives the generation of X-rays, their properties as well as the elements of radiation protection. The utilization characteristics of these radiations in different fields are finally given. (author)

  7. Chest X-Ray

    Medline Plus

    Full Text Available ... by Image/Video Gallery Your Radiologist Explains Chest X-ray Transcript Welcome to Radiology Info dot org! Hello, ... d like to talk with you about chest radiography also known as chest x-rays. Chest x- ...

  8. Microscopic study of dental hard tissues in primary teeth with Dentinogenesis Imperfecta Type II: Correlation of 3D imaging using X-ray microtomography and polarising microscopy.

    Science.gov (United States)

    Davis, Graham R; Fearne, Janice M; Sabel, Nina; Norén, Jörgen G

    2015-07-01

    The aim of this study was to examine the histological appearance of dental hard tissues in primary teeth from children with DI using conventional polarised light microscopy and correlate that with 3D imaging using X-ray microtomograpy (XMT) to gain a further understanding of the dentine structure of teeth diagnosed with dentinogenesis imperfecta. Undecalcified sections of primary teeth from patients diagnosed with Dentinogenesis Imperfecta Type II were examined using polarised light microscopy. XMT was employed for 3D-imaging and analysis of the dentine. The polarised light microscopy and XMT revealed tubular structures in the dentine seen as vacuoles coinciding with the path of normal dentinal tubules but not continuous tubules. The size of the tubules was close to that of capillaries. The largest tubular structures had a direction corresponding to where the pulp tissue would have been located during primary dentine formation. The dysfunctional mineralisation of the dentine and obliteration of the pulp evidently leaves blood vessels in the dentine which have in the main been tied off and, in the undecalcified sections, appear as vacuoles. Although from radiographs, the pulp in teeth affected by Dentinogenesis Imperfect type II appears to be completely obliterated, a network of interconnected vessels may remain. The presence of large dentinal tubules and blood vessels, or the remnants of blood vessels, could provide a pathway for bacteria from the oral cavity. This might account for why some of these teeth develop periapical abscesses in spite of apparently having no pulp. Copyright © 2015 Elsevier Ltd. All rights reserved.

  9. X-ray lasers

    CERN Document Server

    Elton, Raymond C

    2012-01-01

    The first in its field, this book is both an introduction to x-ray lasers and a how-to guide for specialists. It provides new entrants and others interested in the field with a comprehensive overview and describes useful examples of analysis and experiments as background and guidance for researchers undertaking new laser designs. In one succinct volume, X-Ray Lasers collects the knowledge and experience gained in two decades of x-ray laser development and conveys the exciting challenges and possibilities still to come._Add on for longer version of blurb_M>The reader is first introduced

  10. X-ray apparatus

    International Nuclear Information System (INIS)

    Bernstein, S.; Stagg, L.; Lambert, T.W.; Griswa, P.J.

    1976-01-01

    A patient support system for X-ray equipment in arteriographic studies of the heart is described in detail. The support system has been designed to overcome many of the practical problems encountered in using previous types of arteriographic X-ray equipment. The support system is capable of horizontal movement and, by a series of shafts attached to the main support system, the X-ray source and image intensifier or detector may be rotated through the same angle. The system is highly flexible and details are given of several possible operational modes. (U.K.)

  11. X-ray detector

    International Nuclear Information System (INIS)

    Whetten, N.R.; Houston, J.M.

    1977-01-01

    An ionization chamber for use in determining the spatial distribution of x-ray photons in tomography systems comprises a plurality of substantially parallel, planar anodes separated by parallel, planar cathodes and enclosed in a gas of high atomic weight at a pressure from approximately 10 atmospheres to approximately 50 atmospheres. The cathode and anode structures comprise metals which are substantially opaque to x-ray radiation and thereby tend to reduce the resolution limiting effects of x-ray fluoresence in the gas. In another embodiment of the invention the anodes comprise parallel conductive bars disposed between two planar cathodes. Guard rings eliminate surface leakage currents between adjacent electrodes. 8 figures

  12. X-ray apparatus

    International Nuclear Information System (INIS)

    Grady, J.K.

    1985-01-01

    X-ray apparatus is described which has a shutter between the X-ray source and the patient. The shutter controls the level of radiation to which the patient is exposed instead of merely discontinuing the electric power supplied to the source. When the shutter is opened a radiation sensor senses the level of X-radiation. When a preset quantity of X-radiation has been measured an exposure control closes the shutter. Instead of using the radiation sensor, the integrated power supplied to the anode of the X-ray source may be measured. (author)

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... Resources Professions Site Index A-Z X-ray (Radiography) - Bone Bone x-ray uses a very small ... X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray (radiograph) is a noninvasive medical ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... Professions Site Index A-Z X-ray (Radiography) - Bone Bone x-ray uses a very small dose ... limitations of Bone X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray (radiograph) is ...

  15. X-ray examination apparatus

    NARCIS (Netherlands)

    2000-01-01

    The invention relates to an X-ray apparatus which includes an adjustable X-ray filter. In order to adjust an intensity profile of the X-ray beam, an X-ray absorbing liquid is transported to filter elements of the X-ray filter. Such transport is susceptible to gravitational forces which lead to an

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... News Physician Resources Professions Site Index A-Z X-ray (Radiography) - Bone Bone x-ray uses a very ... of Bone X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray (radiograph) is a noninvasive ...

  17. Bone X-Ray (Radiography)

    Science.gov (United States)

    ... News Physician Resources Professions Site Index A-Z X-ray (Radiography) - Bone Bone x-ray uses a very small ... of Bone X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray (radiograph) is a noninvasive ...

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... News Physician Resources Professions Site Index A-Z X-ray (Radiography) - Bone Bone x-ray uses a very small ... of Bone X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray (radiograph) is a noninvasive ...

  19. Abdomen X-Ray (Radiography)

    Science.gov (United States)

    ... News Physician Resources Professions Site Index A-Z X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small ... of an abdominal x-ray? What is abdominal x-ray? An x-ray (radiograph) is a noninvasive medical ...

  20. 75 FR 27504 - Substantial Product Hazard List: Hand-Held Hair Dryers

    Science.gov (United States)

    2010-05-17

    ... immersion during their use. Section 15(a) of the CPSA defines ``substantial product hazard'' to include, a....'' Hand-held hair dryers routinely contain open-coil heating elements that are, in essence, uninsulated..., bathtub, or lavatory). The proposed rule would define ``hand-held hair dryer'' as ``an electrical...

  1. Chest X-Ray

    Medline Plus

    Full Text Available ... Disorders Video: The Basketball Game: An MRI Story Radiology and You Sponsored by Image/Video Gallery Your Radiologist Explains Chest X-ray Transcript Welcome to Radiology Info dot org! Hello, I’m Dr. Geoffrey ...

  2. Chest X-Ray

    Medline Plus

    Full Text Available ... also be useful to help diagnose and monitor treatment for a variety of lung conditions such as pneumonia, emphysema and cancer. A chest x-ray requires no special preparation. ...

  3. X-ray tubes

    International Nuclear Information System (INIS)

    Young, R.W.

    1979-01-01

    A form of x-ray tube is described which provides satisfactory focussing of the electron beam when the beam extends for several feet from gun to target. Such a tube can be used for computerised tomographic scanning. (UK)

  4. Chest X-Ray

    Medline Plus

    Full Text Available ... breath, persistent cough, fever, chest pain or injury. It may also be useful to help diagnose and ... have some concerns about chest x-rays. However, it’s important to consider the likelihood of benefit to ...

  5. Chest X-Ray

    Medline Plus

    Full Text Available ... An MRI Story Radiology and You Sponsored by Image/Video Gallery Your Radiologist Explains Chest X-ray ... posted: How to Obtain and Share Your Medical Images Movement Disorders Video: The Basketball Game: An MRI ...

  6. Chest X-Ray

    Medline Plus

    Full Text Available ... accurate diagnosis far outweighs any risk. For more information about chest x-rays, visit Radiology Info dot ... Inc. (RSNA). To help ensure current and accurate information, we do not permit copying but encourage linking ...

  7. Chest X-Ray

    Medline Plus

    Full Text Available ... Site Index A-Z Spotlight Recently posted: Pancreatic Cancer The Limitations of Online Dose Calculators Video: The ... of lung conditions such as pneumonia, emphysema and cancer. A chest x-ray requires no special preparation. ...

  8. Chest X-Ray

    Medline Plus

    Full Text Available ... exams and use a very small dose of ionizing radiation to produce pictures of the inside of the ... chest x-ray use a tiny dose of ionizing radiation, the benefit of an accurate diagnosis far outweighs ...

  9. Chest X-Ray

    Medline Plus

    Full Text Available ... However, it’s important to consider the likelihood of benefit to your health. While a chest x-ray use a tiny dose of ionizing radiation, the benefit of an accurate diagnosis far outweighs any risk. ...

  10. Chest X-Ray

    Medline Plus

    Full Text Available ... June is Men's Health Month Recently posted: Pancreatic Cancer The Limitations of Online Dose Calculators Video: The ... of lung conditions such as pneumonia, emphysema and cancer. A chest x-ray requires no special preparation. ...

  11. X-ray sources

    International Nuclear Information System (INIS)

    Masswig, I.

    1986-01-01

    The tkb market survey comparatively evaluates the X-ray sources and replacement tubes for stationary equipment currently available on the German market. It lists the equipment parameters of 235 commercially available X-ray sources and their replacement tubes and gives the criteria for purchase decisions. The survey has been completed with December 1985, and offers good information concerning medical and technical aspects as well as those of safety and maintenance. (orig.) [de

  12. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... of knee x-rays. A portable x-ray machine is a compact apparatus that can be taken ... of the body being examined, an x-ray machine produces a small burst of radiation that passes ...

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... that might interfere with the x-ray images. Women should always inform their physician and x-ray ... Safety page for more information about radiation dose. Women should always inform their physician or x-ray ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... ray examination. X-rays usually have no side effects in the typical diagnostic range for this exam. ... x-rays. A Word About Minimizing Radiation Exposure Special care is taken during x-ray examinations to ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... drawer under the table holds the x-ray film or image recording plate . Sometimes the x-ray ... extended over the patient while an x-ray film holder or image recording plate is placed beneath ...

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... are the limitations of Bone X-ray (Radiography)? What is Bone X-ray (Radiography)? An x-ray ( ... leg (shin), ankle or foot. top of page What are some common uses of the procedure? A ...

  17. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  18. X-Ray Absorption with Transmission X-Ray Microscopes

    NARCIS (Netherlands)

    de Groot, F.M.F.

    2016-01-01

    In this section we focus on the use of transmission X-ray microscopy (TXM) to measure the XAS spectra. In the last decade a range of soft X-ray and hard X-ray TXM microscopes have been developed, allowing the measurement of XAS spectra with 10–100 nm resolution. In the hard X-ray range the TXM

  19. X-ray tube

    International Nuclear Information System (INIS)

    Webley, R.S.

    1975-01-01

    The object of the invention described is to provide an X-ray tube providing a scanned X-ray output which does not require a scanned electron beam. This is obtained by an X-ray tube including an anode which is rotatable about an axis, and a source of a beam of energy, for example an electron beam, arranged to impinge on a surface of the anode to generate X-radiation substantially at the region of incidence on the anode surface. The anode is rotatable about the axis to move the region of incidence over the surface. The anode is so shaped that the rotation causes the region of incidence to move in a predetermined manner relative to fixed parts of the tube so that the generated X-radiation is scanned in a predetermined manner relative to the tube. (UK)

  20. X-ray astronomy

    International Nuclear Information System (INIS)

    Giacconi, R.; Setti, G.

    1980-01-01

    This book contains the lectures, and the most important seminars held at the NATO meeting on X-Ray astronomy in Erice, July 1979. The meeting was an opportune forum to discuss the results of the first 8-months of operation of the X-ray satellite, HEAO-2 (Einstein Observatory) which was launched at the end of 1978. Besides surveying these results, the meeting covered extragalactic astronomy, including the relevant observations obtained in other portions of the electromagnetic spectrum (ultra-violet, optical, infrared and radio). The discussion on galactic X-ray sources essentially covered classical binaries, globular clusters and bursters and its significance to extragalactic sources and to high energy astrophysics was borne in mind. (orig.)

  1. Flash x-ray

    International Nuclear Information System (INIS)

    Johnson, Q.; Pellinen, D.

    1976-01-01

    The complementary techniques of flash x-ray radiography (FXR) and flash x-ray diffraction (FXD) provide access to a unique domain in nondestructive materials testing. FXR is useful in studies of macroscopic properties during extremely short time intervals, and FXD, the newer technique, is used in studies of microscopic properties. Although these techniques are similar in many respects, there are some substantial differences. FXD generally requires low-voltage, line-radiation sources and extremely accurate timing; FXR is usually less demanding. Phenomena which can be profitably studied by FXR often can also be studied by FXD to permit a complete materials characterization

  2. X-ray astronomy

    International Nuclear Information System (INIS)

    Narayanan, M.S.

    1976-01-01

    The deployment of detectors outside the deleterious effects of the atmosphere by sending them in space vehicles, has been explained. This has thrown open the entire spectrum of the electromagnetic and particle radiation to direct observations, thus enlarging the vistas of the field of astronomy and astrophysics. The discovery of strong emitters of X-rays such as SCO X-1, NorX-2, transient sources such as Cen X-2, Cen X-4, Cen X-1, Supernova remnants Tan X-1, etc., are reported. The background of the X-ray spectrum as measured during two rocket flights over Thumba, India is presented. (K.B.)

  3. X-ray masks

    International Nuclear Information System (INIS)

    Greenwood, J.C.; Satchell, D.W.

    1984-01-01

    In semiconductor manufacture, where X-ray irradiation is used, a thin silicon membrane can be used as an X-ray mask. This membrane has areas on which are patterns to define the regions to be irradiated. These regions are of antireflection material. With the thin, in the order of 3 microns, membranes used, fragility is a problem. Hence a number of ribs of silicon are formed integral with the membrane, and which are relatively thick, 5 to 10 microns. The ribs may be formed by localised deeper boron deposition followed by a selective etch. (author)

  4. X-ray detector

    International Nuclear Information System (INIS)

    Houston, J.M.; Whetten, N.R.

    1981-01-01

    An ionization chamber for use in determining the spatial distribution of x-ray photons in tomography systems comprises a plurality of substantially parallel, planar anodes separated by parallel, planar cathodes and enclosed in a gas of high atomic weight at a pressure from approximately 10 atmospheres to approximately 50 atmospheres. The cathode and anode structures comprise metals which are substantially opaque to x-ray radiation and thereby tend to reduce the resolution limiting effects of xray fluoresence in the gas. In another embodiment of the invention the anodes comprise parallel conductive bars disposed between two planar cathodes. Guard rings eliminate surface leakage currents between adjacent electrodes

  5. Subluminous X-ray binaries

    NARCIS (Netherlands)

    Armas Padilla, M.

    2013-01-01

    The discovery of the first X-ray binary, Scorpius X-1, by Giacconi et al. (1962), marked the birth of X-ray astronomy. Following that discovery, many additional X-ray sources where found with the first generation of X-ray rockets and observatories (e.g., UHURU and Einstein). The short-timescale

  6. X-Ray Exam: Pelvis

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Pelvis KidsHealth / For Parents / X-Ray Exam: ... Ray Exam: Hip Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  7. X-Ray Exam: Forearm

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Forearm KidsHealth / For Parents / X-Ray Exam: ... Muscles, and Joints Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  8. X-Ray Exam: Foot

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Foot KidsHealth / For Parents / X-Ray Exam: ... Muscles, and Joints Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  9. X-Ray Exam: Wrist

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Wrist KidsHealth / For Parents / X-Ray Exam: ... Muscles, and Joints Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  10. Thoracic spine x-ray

    Science.gov (United States)

    Vertebral radiography; X-ray - spine; Thoracic x-ray; Spine x-ray; Thoracic spine films; Back films ... There is low radiation exposure. X-rays are monitored and regulated to provide the minimum amount of radiation exposure needed to produce the image. Most ...

  11. X-Ray Exam: Finger

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Finger KidsHealth / For Parents / X-Ray Exam: ... Muscles, and Joints Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  12. CRL X-ray tube

    International Nuclear Information System (INIS)

    Kolchevsky, N.N.; Petrov, P.V.

    2015-01-01

    A novel types of X-ray tubes with refractive lenses are proposed. CRL-R X-ray tube consists of Compound Refractive Lens- CRL and Reflection X-ray tube. CRL acts as X-ray window. CRL-T X-ray consists of CRL and Transmission X-ray tube. CRL acts as target for electron beam. CRL refractive lens acts as filter, collimator, waveguide and focusing lens. Properties and construction of the CRL X-ray tube are discussed. (authors)

  13. X rays and condensed matter

    International Nuclear Information System (INIS)

    Daillant, J.

    1997-01-01

    After a historical review of the discovery and study of X rays, the various interaction processes between X rays and matter are described: Thomson scattering, Compton scattering, X-photon absorption through photoelectric effect, and magnetic scattering. X ray sources such as the European Synchrotron Radiation Facility (ESRF) are described. The various X-ray applications are presented: imagery such as X tomography, X microscopy, phase contrast; X-ray photoelectron spectroscopy and X-ray absorption spectroscopy; X-ray scattering and diffraction techniques

  14. 78 FR 27441 - NIJ Evaluation of Hand-Held Cell Phone Detector Devices

    Science.gov (United States)

    2013-05-10

    ...The National Institute of Justice (NIJ) is soliciting interest in supplying hand-held cell phone detector devices for participation in an evaluation by the NIJ Corrections Technology Center of Excellence (CXCoE).

  15. X-ray beam generator

    International Nuclear Information System (INIS)

    Koller, T.J.; Randmer, J.A.

    1977-01-01

    A method of minimizing the preferential angular absorption of the divergent beam from an X-ray generator is described. The generator consists of an X-ray shielded housing with an X-ray transmissive window symmetrically placed in radial alignment with a focal spot area on a sloped target surface of an X-ray tube in the housing. The X-ray tube may be of the stationary anode type or of the rotating anode type. (U.K.)

  16. Chest X-Ray

    Medline Plus

    Full Text Available Toggle navigation Test/Treatment Patient Type Screening/Wellness Disease/Condition Safety En Español More Info Images/Videos ... x-ray is used to evaluate the lungs, heart and chest wall and may be used to ...

  17. X-ray sources

    International Nuclear Information System (INIS)

    Bonse, U.

    1979-11-01

    The author describes several possibilities for the production of X-radiation. Especially he discusses the use of bremsstrahlung at electron impact on solid targets and the synchrotron radiation. He presents some equations for the calculation of X-ray intensities. Especially the X-radiation from the DORIS storage ring is discussed. (HSI)

  18. Medical x-ray

    International Nuclear Information System (INIS)

    Abd Aziz Mhd Ramli; Gui Ah Auu; Husaini Salleh; Idris Besar; Mohd Ashhar Khalid; Muhammad Jamal Md Isa; Shaharuddin Mohd; Siti Najila Mohd Janib; Mohamed Ali Abdul Khader; Mahalatchimi Dave; Mohd Fazly Abdul Rahim; Ng Chee Moon; Ram Piari; Teoh Hoon Heng; Lee Peter

    2004-01-01

    This book describes the fundamental subject about medical radiography. It is a multidisciplinary field that requires cross professional input from scientists, engineers and medical doctors. However, it is presented in simple language to suit different levels of readers from x-ray operators and radiographers to physists, general practitioners and radiology specialists.The book is written in accordance to the requirements of the standard syllabus approved by the Ministry of Health Malaysia for the training of medical x-ray operator and general practitioners. In general, the content is not only designed to provide relevant and essential subject for related professionals in medical radiological services such as x-ray operator, radiographer and radiologists, but also to address those in associated radiological services including nurses, medical technologists and physicists.The book is organized and arranged sequentially into 3 parts for easy reference: Radiation safety; X-ray equipment and associated facilities; Radiography practices. With proper grasping of all these parts, the radiological services could be provided with confident and the highest professional standard. Thus, medical imaging with highest quality that can provide useful diagnostic information at minimum doses and at cost effective could be assured

  19. Chest X-Ray

    Medline Plus

    Full Text Available ... change into a gown. You may have some concerns about chest x-rays. However, it’s important to ... You Sponsored by About Us | Contact Us | FAQ | Privacy | Terms of Use | Links | Site Map Copyright © 2018 ...

  20. Cross-sectional imaging with rotational panoramic X-ray machine for preoperative assessment of dental implant site. Comparisons of imaging properties with conventional film tomography and computed tomography

    International Nuclear Information System (INIS)

    Makihara, Masahiro; Nishikawa, Keiichi; Kuroyanagi, Kinya

    2001-01-01

    To clarify the validity of cross-sectional imaging with rotational panoramic x-ray machine for preoperative assessment of the dental implant site, the imaging properties were compared with those of spiral tomography and multi-planer reconstruction (MPR) manipulation of x-ray computed tomography. Cross-sectional imaging of the maxilla and mandible of an edentulous dry skull was performed by each technique at an image layer thickness of 1 mm. Steel spheres were used to identify cross-sectional planes and measure distance. Six oral radiologists scored the image clarity of structures with 5-grade rating scales and measured the distance between images of 2 steel spheres. Each measured distance was divided by the magnification factor. The actual distance was also measured on the skull. The score and the distance were statistically compared. The Spearman's rank correlation coefficients for the score and the absolute values of the difference in distances measured by different observers were calculated as test units to compare inter-observer agreements statistically. The same observation and measurement were repeated to compare intra-observer agreement. Image clarity of the linear tomography available with a panoramic machine was comparable to spiral tomography and superior to MPR, except for the cortical bone on the lingual side. The inter- and intra-observer agreements were comparable. The accuracy for measurement of distance, the inter- and intra-observer agreements were also comparable to the spiral tomography and superior to those of MPR. Therefore, it is concluded that cross-sectional imaging with a rotational panoramic x-ray machine is useful for preoperative assessment of the dental implant site. (author)

  1. A differential scanning calorimetric and X-ray diffraction

    Indian Academy of Sciences (India)

    X-ray diffraction analysis of the two groups demonstrated predominance of austenitic ... Discrete crystallographic structure and absence of multiple phases showed ... Division of Orthodontics, Indian Army, 10 Corps Dental Unit, C/O 56 APO, ...

  2. X ray Production. Chapter 5

    Energy Technology Data Exchange (ETDEWEB)

    Nowotny, R. [Medical University of Vienna, Vienna (Austria)

    2014-09-15

    The differential absorption of X rays in tissues and organs, owing to their atomic composition, is the basis for the various imaging methods used in diagnostic radiology. The principles in the production of X rays have remained the same since their discovery. However, much refinement has gone into the design of X ray tubes to achieve the performance required for today’s radiological examinations. In this chapter, an outline of the principles of X ray production and a characterization of the radiation output of X ray tubes will be given. The basic processes producing X rays are dealt with in Section 1.4.

  3. X-ray filter for x-ray powder diffraction

    Science.gov (United States)

    Sinsheimer, John Jay; Conley, Raymond P.; Bouet, Nathalie C. D.; Dooryhee, Eric; Ghose, Sanjit

    2018-01-23

    Technologies are described for apparatus, methods and systems effective for filtering. The filters may comprise a first plate. The first plate may include an x-ray absorbing material and walls defining first slits. The first slits may include arc shaped openings through the first plate. The walls of the first plate may be configured to absorb at least some of first x-rays when the first x-rays are incident on the x-ray absorbing material, and to output second x-rays. The filters may comprise a second plate spaced from the first plate. The second plate may include the x-ray absorbing material and walls defining second slits. The second slits may include arc shaped openings through the second plate. The walls of the second plate may be configured to absorb at least some of second x-rays and to output third x-rays.

  4. X-ray refractometer

    International Nuclear Information System (INIS)

    Tur'yanskij, A.G.; Pirshin, I.V.

    2001-01-01

    Paper introduces a new circuit of X-ray refractometer to study angular and spectral features of refracted radiation within hard X-ray range. Refractometer incorporates two goniometers, two crystal-analyzers and three radiation detectors. The maximum distance between radiation source focal point and a receiving slit of the second goniometer is equal to 1.4 m. For the first time one obtained refraction patterns of fine-film specimens including C/Si stressed structure. Paper describes a new technique of refractometry via specimen oscillation at fixed position of a detecting device. Paper presents the measurement results of oscillation refraction patterns for specimens of melted quartz and ZnSe single crystal [ru

  5. X-ray radiotherapy

    International Nuclear Information System (INIS)

    Tronc, D.

    1995-01-01

    Full text: The most common form of radio therapy is X-ray therapy, where a beam of photons or their parent electrons break down hydrogen bonds within the body's cells and remove certain DNA information necessary for cell multiplication. This process can eradicate malignant cells leading to complete recovery, to the remission of some cancers, or at least to a degree of pain relief. The radiotherapy instrument is usually an electron linac, and the electrons are used either directly in 'electrotherapy' for some 10% of patients, or the electrons bombard a conversion target creating a broad beam of high energy photons or 'penetration X-rays'. The simplest machine consists of several accelerating sections at around 3 GHz, accelerating electrons to 6 MeV; a cooled tungsten target is used to produce a 4 Gray/min X-ray field which can be collimated into a rectangular shape at the patient position. This tiny linac is mounted inside a rotating isocentric gantry above the patient who must remain perfectly still. Several convergent beams can also be used to increase the delivered dose. More sophisticated accelerators operate at up to 18 MeV to increase penetration depths and decrease skin exposure. Alternatively, electrotherapy can be used with different energies for lower and variable penetration depths - approximately 0.5 cm per MeV. In this way surface tissue may be treated without affecting deeper and more critical anatomical regions. This type of linac, 1 to 2 metres long, is mounted parallel to the patient with a bending magnet to direct the beam to the radiotherapy system, which includes the target, thick movable collimator jaws, a beam field equalizer, dose rate and optical field simulation and energy controls. There are over 2000 acceleratorbased X-ray treatment units worldwide. Western countries have up to two units per million population, whereas in developing countries such as Bangladesh, the density is only one per 100 million. Several

  6. X-ray generator

    Energy Technology Data Exchange (ETDEWEB)

    Zucarias, A; Shepherd, J W

    1982-09-08

    An X-ray tube has a tubular envelope with a cathode for directing an electron beam onto a focal spot area of a spaced anode target to generate X-rays. The target is mounted for axial rotation on one end of a rotor disposed in an end portion of the envelope and encircled by a stator of an alternating current induction motor. An annular shield of high permeability magnetic material extends transversely between the electron beam and the stator of the induction motor for shunting stray or fringe electromagnetic fields established by the stator away from the electron beam to avoid consequent lateral deflections of the electron and corresponding lateral movements of the focal spot area.

  7. X-ray microtomography

    International Nuclear Information System (INIS)

    Dunsmuir, J.H.; Ferguson, S.R.; D'Amico, K.L.; Stokes, J.P.

    1991-01-01

    In this paper the authors describe the application of a new high-resolution X-ray tomographic microscope to the study of porous media. The microscope was designed to exploit the properties of a synchrotron X-ray source to perform three dimensional tomography on millimeter sized objects with micron resolution and has been used in materials science studies with both synchrotron and conventional and synchrotron sources will be compared. In this work the authors have applied the microscope to measure the three dimensional structure of fused bead packs and berea sandstones with micron resolution and have performed preliminary studies of flow in these media with the microscope operated in a digital subtraction radiography mode. Computer graphics techniques have been applied to the data to visually display the structure of the pore body system. Tomographic imaging after flow experiments should detect the structure of the oil-water interface in the pore network and this work is ongoing

  8. X-ray diffraction

    International Nuclear Information System (INIS)

    Einstein, J.R.; Wei, C.H.

    1982-01-01

    We have been interested in structural elucidation by x-ray diffraction of compounds of biological interest. Understanding exactly how atoms are arranged in three-dimensional arrays as molecules can help explain the relationship between structure and functions. The species investigated may vary in size and shape; our recent studies included such diverse substances as antischistosomal drugs, a complex of cadmium with nucleic acid base, nitrate salts of adenine, and proteins

  9. X-ray apparatus

    International Nuclear Information System (INIS)

    Tomita, Chuji.

    1980-01-01

    A principal object of the present invention is to provide an X-ray apparatus which is such that the distance between the surface of the patient's table and the floor on which the apparatus is installed is sufficiently small in the horizontal position of the patient's table of the roentgenographical pedestal and that the rotation of the pedestal from the horizontal position to a tilted position and further to the vertical position of the table can be carried out smoothly. (auth)

  10. X-ray Ordinance

    International Nuclear Information System (INIS)

    Kramer, R.; Zerlett, G.

    1983-01-01

    This commentary, presented as volume 2 of the Deutsches Strahlenschutzrecht (German legislation on radiation protection) deals with the legal provisions of the ordinance on the protection against harmful effects of X-radiation (X-ray Ordinance - RoeV), of March 1, 1973 (announced in BGBl.I, page 173), as amended by the ordinance on the protection against harmful effects of ionizing radiation, of October 13, 1976 (announced in BGBl. I, page 2905). Thus volume 2 completes the task started with volume 1, namely to present a comprehensive view and account of the body of laws governing radiation protection, a task which was thought useful as developments in the FRG led to regulations being split up into the X-ray Ordinance, and the Radiation Protection Ordinance. In order to present a well-balanced commentary on the X-ray Ordinance, it was necessary to discuss the provisions both from the legal and the medical point of view. This edition takes into account the Fourth Public Notice of the BMA (Fed. Min. of Labour and Social Affairs) concerning the implementation of the X-ray Ordinance of January 4, 1982, as well as court decisions and literature published in this field, until September 1982. In addition, the judgment of the Federal Constitutional Court, dated October 19, 1982, concerning the voidness of the law on government liability, and two decisions by the Federal High Court, dated November 23, 1982, concerning the right to have insight into medical reports - of great significance in practice - have been considered. This commentary therefore is up to date with current developments. (orig.) [de

  11. Producing x-rays

    International Nuclear Information System (INIS)

    Mallozzi, P.J.; Epstein, H.M.; Jung, R.G.; Applebaum, D.C.; Fairand, B.P.; Gallagher, W.J.

    1977-01-01

    A method of producing x-rays by directing radiant energy from a laser onto a target is described. Conversion efficiency of at least about 3 percent is obtained by providing the radiant energy in a low-power precursor pulse of approximately uniform effective intensity focused onto the surface of the target for about 1 to 30 nanoseconds so as to generate an expanding unconfined coronal plasma having less than normal solid density throughout and comprising a low-density (underdense) region wherein the plasma frequency is less than the laser radiation frequency and a higher-density (overdense) region wherein the plasma frequency is greater than the laser radiation frequency and, about 1 to 30 nanoseconds after the precursor pulse strikes the target, a higher-power main pulse focused onto the plasma for about 10 -3 to 30 nanoseconds and having such power density and total energy that the radiant energy is absorbed in the underdense region and conducted into the overdense region to heat it and thus to produce x-rays therefrom with the plasma remaining substantially below normal solid density and thus facilitating the substantial emission of x-rays in the form of spectral lines arising from nonequilibrium ionization states

  12. Ultrasonography with a hand-held device for the diagnosis of acute appendicitis

    International Nuclear Information System (INIS)

    Kameda, Toru; Takahashi, Isao

    2009-01-01

    The purpose of this study was to evaluate the accuracy of ultrasonography (US) with a hand-held device for the diagnosis of acute appendicitis in the emergency room. US with a hand-held device was performed by the first author in 33 patients suspected of having appendicitis in the emergency room. From these 33 patients, 24 who subsequently underwent computed tomography (CT) or surgery were included in this study. The accuracy of US with the hand-held device for the diagnosis of acute appendicitis was evaluated based on the findings of CT or surgery. CT and surgery were performed in 22 and 12 patients, respectively. Final diagnoses were acute appendicitis (n=18), terminal ileitis (n=2), pelvic inflammatory disease (n=2), diverticulitis (n=1), and ureterolithiasis (n=1). The US yielded a sensitivity of 78% and a positive predictive value of 100%. The shortest distance between the abdominal wall and the appendix measured on CT was less than 40 mm in 11 patients. In ten (91%) of the 11 patients US with the hand-held device showed the swollen appendix. US with a hand-held device is potentially useful in the positive identification of acute appendicitis, but further investigation is needed to prove its utility in the routine diagnosis of acute appendicitis. (author)

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... the body. X-rays are the oldest and most frequently used form of medical imaging. A bone ... bones. top of page How should I prepare? Most bone x-rays require no special preparation. You ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... evaluation. National and international radiology protection organizations continually review and update the technique standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... can be taken to the patient in a hospital bed or the emergency room. The x-ray ... position possible that still ensures x-ray image quality. top of page Who interprets the results and ...

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... position possible that still ensures x-ray image quality. top of page Who interprets the results and ...

  17. X-ray detector array

    International Nuclear Information System (INIS)

    Houston, J.M.

    1980-01-01

    The object of the invention (an ionization chamber X-ray detector array for use with high speed computerised tomographic imaging apparatus) is to reduce the time required to produce a tomographic image. The detector array described determines the distribution of X-ray intensities in one or more flat, coplanar X-ray beams. It comprises three flat anode sheets parallel to the X-ray beam, a plurality of rod-like cathodes between the anodes, a detector gas between the electrodes and a means for applying a potential between the electrodes. Each of the X-ray sources is collimated to give a narrow, planar section of X-ray photons. Sets of X-ray sources in the array are pulsed simultaneously to obtain X-ray transmission data for tomographic image reconstruction. (U.K.)

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... tissue shows up in shades of gray and air appears black. Until recently, x-ray images were ... position possible that still ensures x-ray image quality. top of page Who interprets the results and ...

  19. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... changes seen in metabolic conditions. assist in the detection and diagnosis of bone cancer . locate foreign objects ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  20. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that ...

  1. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... up in shades of gray and air appears black. Until recently, x-ray images were maintained on ... Safety page for more information about radiation dose. Women should always inform their physician or x-ray ...

  2. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... wrist, arm, elbow, shoulder, spine, pelvis, hip, thigh, knee, leg (shin), ankle or foot. top of page ... the patient standing upright, as in cases of knee x-rays. A portable x-ray machine is ...

  3. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures ...

  4. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... conditions. Imaging with x-rays involves exposing a part of the body to a small dose of ... body. Once it is carefully aimed at the part of the body being examined, an x-ray ...

  5. X-Ray Exam: Hip

    Science.gov (United States)

    ... for Educators Search English Español X-Ray Exam: Hip KidsHealth / For Parents / X-Ray Exam: Hip What's in this article? What It Is Why ... You Have Questions Print What It Is A hip X-ray is a safe and painless test ...

  6. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... X-rays are a form of radiation like light or radio waves. X-rays pass through most objects, including the body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small ...

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... x-rays. top of page What does the equipment look like? The equipment typically used for bone x-rays consists of ... and joint abnormalities, such as arthritis. X-ray equipment is relatively inexpensive and widely available in emergency ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... The x-ray tube is connected to a flexible arm that is extended over the patient while an x-ray film holder or image recording plate is placed beneath the patient. top of page How does the procedure work? X-rays are a form of radiation like ...

  9. X-Ray Exam: Ankle

    Science.gov (United States)

    ... for Educators Search English Español X-Ray Exam: Ankle KidsHealth / For Parents / X-Ray Exam: Ankle What's in this article? What It Is Why ... You Have Questions Print What It Is An ankle X-ray is a safe and painless test ...

  10. Driver hand-held cellular phone use: a four-year analysis.

    Science.gov (United States)

    Eby, David W; Vivoda, Jonathon M; St Louis, Renée M

    2006-01-01

    The use of hand-held cellular (mobile) phones while driving has stirred more debate, passion, and research than perhaps any other traffic safety issue in the past several years. There is ample research showing that the use of either hand-held or hands-free cellular phones can lead to unsafe driving patterns. Whether or not these performance deficits increase the risk of crash is difficult to establish, but recent studies are beginning to suggest that cellular phone use elevates crash risk. The purpose of this study was to assess changes in the rate of hand-held cellular phone use by motor-vehicle drivers on a statewide level in Michigan. This study presents the results of 13 statewide surveys of cellular phone use over a 4-year period. Hand-held cellular phone use data were collected through direct observation while vehicles were stopped at intersections and freeway exit ramps. Data were weighted to be representative of all drivers traveling during daylight hours in Michigan. The study found that driver hand-held cellular phone use has more than doubled between 2001 and 2005, from 2.7% to 5.8%. This change represents an average increase of 0.78 percentage points per year. The 5.8% use rate observed in 2005 means that at any given daylight hour, around 36,550 drivers were conversing on cellular phones while driving on Michigan roadways. The trend line fitted to these data predicts that by the year 2010, driver hand-held cellular phone use will be around 8.6%, or 55,000 drivers at any given daylight hour. These results make it clear that cellular phone use while driving will continue to be an important traffic safety issue, and highlight the importance of continued attempts to generate new ways of alleviating this potential hazard.

  11. SMM x ray polychromator

    Science.gov (United States)

    Saba, J. L. R.

    1993-01-01

    The objective of the X-ray Polychromator (XRP) experiment was to study the physical properties of solar flare plasma and its relation to the parent active region to understand better the flare mechanism and related solar activity. Observations were made to determine the temperature, density, and dynamic structure of the pre-flare and flare plasma as a function of wavelength, space and time, the extent to which the flare plasma departs from thermal equilibrium, and the variation of this departure with time. The experiment also determines the temperature and density structure of active regions and flare-induced changes in the regions.

  12. Obstetric X-rays

    International Nuclear Information System (INIS)

    Mwachi, M.K.

    2006-01-01

    Radiography of the pelvis should never be taken to diagnose early pregnancy, because of potential hazards of radiation damage to the growing foetus. the only indication occurs in the last week of pregnancy (37 weeks). Obstetric X-ray will help you answer like confirmation of malposition,multiple pregnancies; fetal abnormalities e.g. hydrocephalus, foetal disposition. The choice of radiographic projection will help give foetal presentation, disposition as well as foetal maturity. The search pattern helps you determine maternal and spine deformity, foetal spine and head , foetal presentation and any other anomalies

  13. X-ray film

    International Nuclear Information System (INIS)

    Arndt, U.W.; Gilmore, D.J.; Wonacott, A.J.

    1977-01-01

    The performance of film as an X-ray detector is discussed and its behaviour is compared with that of a perfect Poissonian detector. The efficiency of microdensitometry as a method of extracting the information recorded on the film is discussed. More emphasis is placed in the precision of microdensitometric measurements than on the more obvious characteristic of film speed. The effects of chemical fog and background on the precision of the measurements is considered and it is concluded that the final limit to precision is set by the chemical fog. (B.D.)

  14. X-ray diffraction

    International Nuclear Information System (INIS)

    Vries, J.L. de.

    1976-01-01

    The seventh edition of Philips' Review of literature on X-ray diffraction begins with a list of conference proceedings on the subject, organised by the Philips' organisation at regular intervals in various European countries. This is followed by a list of bulletins. The bibliography is divided according to the equipment (cameras, diffractometers, monochromators) and its applications. The applications are subdivided into sections for high/low temperature and pressure, effects due to the equipment, small angle scattering and a part for stress, texture and phase analyses of metals and quantitative analysis of minerals

  15. Fabrication of applicator system of miniature X-ray tube based on carbon nanotubes for a skin cancer therapy

    International Nuclear Information System (INIS)

    Park, Han Beom; Kim, Hyun Jin; Lee, Ju Hyuk; Ha, Jun Mok; Cho, Sung Oh

    2016-01-01

    A miniature X-ray tube is a small X-ray generation device generally with a diameter of less than 10 mm. Because of the feasible installation in a spatially constrained area and the possibility of electrical on/off control, miniature X-ray tubes can be widely used for nondestructive X-ray radiography, hand held X-ray spectrometers, electric brachytherapy, and interstitial or intracavitary radiation therapy or imaging with the substitution of radioactive isotopes. Miniature X-ray tubes have been developed mostly using thermionic electron sources or secondary X-ray emission. The X-ray tube show excellent field emission properties and good X-ray spectrum. Also, the flattening filter was made to irradiate uniformly. The X-ray dose radial uniformities between installed flattening filter and non-installed flattening filter were measured. When flattening filter is equipped, X-ray uniformity was improved from higher than 20% to lower than 10%. As a result, the fabricated applicator system of the miniature X-ray tube using optimized flattening filter exhibited fairly excellent properties

  16. A comparative transmission electron microscopy, energy dispersive x-ray spectroscopy and spatially resolved micropillar compression study of the yttria partially stabilised zirconia - porcelain interface in dental prosthesis

    Energy Technology Data Exchange (ETDEWEB)

    Lunt, Alexander J.G., E-mail: alexander.lunt@chch.ox.ac.uk [Department of Engineering Science, University of Oxford, Parks Road, Oxford, Oxfordshire OX1 3PJ (United Kingdom); Mohanty, Gaurav, E-mail: gaurav.mohanty@empa.ch [EMPA Materials Science & Technology, Feuerwerkerstrasse 39, CH-3602 Thun (Switzerland); Ying, Siqi, E-mail: siqi.ying@eng.ox.ac.uk [Department of Engineering Science, University of Oxford, Parks Road, Oxford, Oxfordshire OX1 3PJ (United Kingdom); Dluhoš, Jiří, E-mail: jiri.dluhos@tescan.cz [TESCAN Brno, s.r.o., Libušina tř. 1, 623 00 Brno-Kohoutovice (Czech Republic); Sui, Tan, E-mail: tan.sui@eng.ox.ac.uk [Department of Engineering Science, University of Oxford, Parks Road, Oxford, Oxfordshire OX1 3PJ (United Kingdom); Neo, Tee K., E-mail: neophyte@singnet.com.sg [Specialist Dental Group, Mount Elizabeth Orchard, 3 Mount Elizabeth, #08-03/08-08/08-10, 228510 (Singapore); Michler, Johann, E-mail: johann.michler@empa.ch [EMPA Materials Science & Technology, Feuerwerkerstrasse 39, CH-3602 Thun (Switzerland); Korsunsky, Alexander M., E-mail: alexander.korsunsky@eng.ox.ac.uk [Department of Engineering Science, University of Oxford, Parks Road, Oxford, Oxfordshire OX1 3PJ (United Kingdom)

    2015-12-01

    Recent studies into the origins of failure of yttria partially stabilised zirconia–porcelain veneered prosthesis have revealed the importance of micro-to-nano scale characterisation of this interface zone. Current understanding suggests that the heat treatment, residual stresses and varying microstructure at this location may contribute to near-interface porcelain chipping. In this study the chemical, microstructural and mechanical property variation across the interfacial zone has been characterised at two differing length scales and using three independent techniques; energy dispersive X-ray spectroscopy, transmission electron microscopy and micropillar compression. Energy dispersive X-ray spectroscopy mapping of the near-interface region revealed, for the first time, that the diffusional lengths of twelve principal elements are limited to within 2–6 μm of the interface. This study also revealed that 0.2–2 μm diameter zirconia grains had become detached from the bulk and were embedded in the near-interface porcelain. Transmission electron microscopy analysis demonstrated the presence of nanoscale spherical features, indicative of tensile creep induced voiding, within the first 0.4–1.5 μm from the interface. Within zirconia, variations in grain size and atomistic structure were also observed within the 3 μm closest to the interface. Micropillar compression was performed over a 100 μm range on either side of the interface at the spatial resolution of 5 μm. This revealed an increase in zirconia and porcelain loading modulus at close proximities (< 5 μm) to the interface and a decrease in zirconia modulus at distances between 6 and 41 μm from this location. The combination of the three experimental techniques has revealed intricate details of the microstructural, chemical and consequently mechanical heterogeneities in the YPSZ–porcelain interface, and demonstrated that the length scales typically associated with this behaviour are approximately ± 5

  17. Soft x-ray lasers

    International Nuclear Information System (INIS)

    Matthews, D.L.; Rosen, M.D.

    1988-01-01

    One of the elusive dreams of laser physicists has been the development of an x-ray laser. After 25 years of waiting, the x-ray laser has at last entered the scientific scene, although those now in operation are still laboratory prototypes. They produce soft x rays down to about five nanometers. X-ray lasers retain the usual characteristics of their optical counterparts: a very tight beam, spatial and temporal coherence, and extreme brightness. Present x-ray lasers are nearly 100 times brighter that the next most powerful x-ray source in the world: the electron synchrotron. Although Lawrence Livermore National Laboratory (LLNL) is widely known for its hard-x-ray laser program which has potential applications in the Strategic Defense Initiative, the soft x-ray lasers have no direct military applications. These lasers, and the scientific tools that result from their development, may one day have a place in the design and diagnosis of both laser fusion and hard x-ray lasers. The soft x-ray lasers now in operation at the LLNL have shown great promise but are still in the primitive state. Once x-ray lasers become reliable, efficient, and economical, they will have several important applications. Chief among them might be the creation of holograms of microscopic biological structures too small to be investigated with visible light. 5 figs

  18. 21 CFR 872.1810 - Intraoral source x-ray system.

    Science.gov (United States)

    2010-04-01

    ... (CONTINUED) MEDICAL DEVICES DENTAL DEVICES Diagnostic Devices § 872.1810 Intraoral source x-ray system. (a... structures. The x-ray source (a tube) is located inside the mouth. This generic type of device may include... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Intraoral source x-ray system. 872.1810 Section...

  19. X-ray spectrometry

    International Nuclear Information System (INIS)

    Markowicz, A.A.; Van Grieken, R.E.

    1986-01-01

    In the period under review, i.e, through 1984 and 1985, some 600 articles on XRS (X-ray spectrometry) were published; most of these have been scanned and the most fundamental ones are discussed. All references will refer to English-language articles, unless states otherwise. Also general books have appeared on quantitative EPXMA (electron-probe X-ray microanalysis) and analytical electron microscopy (AEM) as well as an extensive review on the application of XRS to trace analysis of environmental samples. In the period under review no radically new developments have been seen in XRS. However, significant improvements have been made. Gain in intensities has been achieved by more efficient excitation, higher reflectivity of dispersing media, and better geometry. Better understanding of the physical process of photon- and electron-specimen interactions led to complex but more accurate equations for correction of various interelement effects. Extensive use of micro- and minicomputers now enables fully automatic operation, including qualitative analysis. However, sample preparation and presentation still put a limit to further progress. Although some authors find XRS in the phase of stabilization or even stagnation, further gradual developments are expected, particularly toward more dedicated equipment, advanced automation, and image analysis systems. Ways are outlined in which XRS has been improved in the 2 last years by excitation, detection, instrumental, methodological, and theoretical advances. 340 references

  20. Mechanism of Action of TiF4 on Dental Enamel Surface: SEM/EDX, KOH-Soluble F, and X-Ray Diffraction Analysis.

    Science.gov (United States)

    Comar, Lívia P; Souza, Beatriz M; Al-Ahj, Luana P; Martins, Jessica; Grizzo, Larissa T; Piasentim, Isabelle S; Rios, Daniela; Buzalaf, Marília Afonso Rabelo; Magalhães, Ana Carolina

    2017-10-12

    This in vitro study aimed to evaluate the action of TiF4 on sound and carious bovine and human enamel. Sound (S) and pre-demineralised (DE) bovine and human (primary and permanent) enamel samples were treated with TiF4 (pH 1.0) or NaF varnishes (pH 5.0), containing 0.95, 1.95, or 2.45% F for 12 h. The enamel surfaces were analysed using SEM-EDX (scanning electron microscopy/energy-dispersive X-ray spectroscopy) (n = 10, 5 S and 5 DE) and KOH-soluble fluoride was quantified (n = 20, 10 S and 10 DE). Hydroxyapatite powder produced by precipitation method was treated with the corresponding fluoride solutions for 1 min (n = 2). The formed compounds were detected using X-ray diffraction (XRD). All TiF4 varnishes produced a coating layer rich in Ti and F on all types of enamel surface, with micro-cracks in its extension. TiF4 (1.95 and 2.45% F) provided higher fluoride deposition than NaF, especially for bovine enamel (p enamel. The Ti content was higher for bovine and human primary enamel than human permanent enamel, with some differences between S and DE. The XRD analysis showed that TiF4 induced the formation of new compounds such as CaF2, TiO2, and Ti(HPO4)2·H2O. In conclusion, TiF4 (>0.95% F) interacts better, when compared to NaF, with bovine and human primary enamel than with human permanent enamel. TiF4 provoked higher F deposition compared to NaF. Carious enamel showed higher F uptake than sound enamel by TiF4 application, while Ti uptake was dependent on the enamel condition and origin. © 2017 S. Karger AG, Basel.

  1. Efficacy of pumping manipulation treatment for closed lock of the temporomandibular joint disorder. Correlation between arthrographic findings using limited cone beam X-ray CT for dental use and mouth opening distance in 20 cases

    International Nuclear Information System (INIS)

    Honda, Kazuya; Uehara, Tamotsu; Arai, Yoshinori; Kashima, Masahiro; Tsukimura, Naoki; Honda, Masahiko; Iwai, Kazuo; Terakado, Masaaki; Shinoda, Koji

    2002-01-01

    A study was conducted to evaluate the efficacy of pumping manipulation treatment for closed lock of the temporomandibular joint (TMJ) disorder using limited cone beam X-ray CT for dental use. The subjects were 20 patients with TMJ closed lock. Arthrography and pumping manipulation treatment were performed, and the correlation between maximal mouth opening and arthrographic findings was examined. Arthrography showed 16 cases of anterior disk displacement, and 4 cases of sideways displacement. Disk configuration showed 15 abnormal cases and 3 cases of disk perforation. Before treatment, mouth opening distance was 24.2 mm and 1 week after treatment it was 34.4 mm. After 3 months this had improved significantly to 41.0 mm. Comparison of mouth opening distance with arthrographic findings showed that disk perforation was significantly different after 3 months. These results suggest that pumping manipulation treatment might be useful in patients with TMJ closed lock without internal derangement or disk perforation. (author)

  2. Development of a dosimetric system for dental X-ray equipment quality control; Desenvolvimento de um sistema dosimetrico para controle de qualidade de equipamentos de raios-X odontologicos

    Energy Technology Data Exchange (ETDEWEB)

    Melo, Francisco Almeida de

    2002-08-01

    An electronic instrument with digital readout was designed and constructed to provide fast, simple and non-invasive measurements of X-ray dental equipment parameters. This instrument is capable of evaluating the entrance dose, exposure time, tube voltage (kVp) and beam filtration. It consists of a set of five photodiodes connected each one in the photocurrent mode to the input of a designed integrating electrometer. Three of the detectors are fixed under aluminium filters with different thicknesses, one is fixed under a 0.3 mm copper filter and the other has no filtration. The readings of the three detectors under aluminium filters and of the bare detector permit the determination of the half-value layer, which is used to calculate the beam filtration. The ratio between the readings of the detector below the copper filter and the one without filtration is used to determine the tube voltage. The signal produced by the detector without filter is used to evaluate the patient entrance dose, and to active an electronic timer for measuring the real exposure time. The tests and calibration of the instrument in different voltages in the 59 - 70kVp range, showed that its response is both stable and reproducible to within 1%. The instrument response was compared to the one from a commercial non-invasive X-ray test equipment (Gammex RMI Multifunction kVp meter). The results showed that the response of the developed instrument is in good agreement with the RMI meter which is the standard equipment for such measurements. These results indicate that the dosimetric system is suitable for use in Dental Quality Assurance Programs. (author)

  3. How to use hand-held computers to evaluate wood drying.

    Science.gov (United States)

    Howard N. Rosen; Darrell S. Martin

    1985-01-01

    Techniques have been developed to evaluate end generate wood drying curves with hand-held computers (3-5K memory). Predictions of time to dry to a specific moisture content, drying rates, and other characteristics of wood drying curves can be made. The paper describes the development of programs and illustrates their use.

  4. 78 FR 73415 - Safety Standard for Hand-Held Infant Carriers

    Science.gov (United States)

    2013-12-06

    ... modifications to the ASTM test procedure set forth in ASTM F2050-12 so that the test produces more repeatable... skills necessary for the preparation of reports or records; and a description of the steps the agency has... the Rule on Small Business There are at least 47 firms currently known to be marketing hand- held...

  5. Survey reveals public open to ban on hand-held cell phone use and texting.

    Science.gov (United States)

    2013-01-01

    A study performed by the Bureau of Transportation Statistics (BTS) reveals that the public is open to a ban on hand-held cell phone use while driving. The study is based on data from 2009s Omnibus Household Survey (OHS), which is administered by B...

  6. Applying Hand-Held 3D Printing Technology to the Teaching of VSEPR Theory

    Science.gov (United States)

    Dean, Natalie L.; Ewan, Corrina; McIndoe, J. Scott

    2016-01-01

    The use of hand-held 3D printing technology provides a unique and engaging approach to learning VSEPR theory by enabling students to draw three-dimensional depictions of different molecular geometries, giving them an appreciation of the shapes of the building blocks of complex molecular structures. Students are provided with 3D printing pens and…

  7. Radiation safety evaluation of a hand-held, battery operated image intensifier

    International Nuclear Information System (INIS)

    Wilson, O.J.; Young, B.F.

    1987-01-01

    A portable, hand-held, fluoroscopic unit intended for medical and industrial use was tested to verify the claim of the manufacturers that the radiation doses to the patient and user are low, and comparable to those received from standard radiographic procedures. The first claim was substantiated but not the second. A number of concerns arising from the use of this unit are discussed

  8. Integrating a Hand Held computer and Stethoscope into a Fetal Monitor

    Science.gov (United States)

    Ahmad Soltani, Mitra

    2009-01-01

    This article presents procedures for modifying a hand held computer or personal digital assistant (PDA) into a versatile device functioning as an electronic stethoscope for fetal monitoring. Along with functioning as an electronic stethoscope, a PDA can provide a useful information source for a medical trainee. Feedback from medical students, residents and interns suggests the device is well accepted by medical trainees. PMID:20165517

  9. The Weak Link HP-41C hand-held calculator program

    Science.gov (United States)

    Ross A. Phillips; Penn A. Peters; Gary D. Falk

    1982-01-01

    The Weak Link hand-held calculator program (HP-41C) quickly analyzes a system for logging production and costs. The production equations model conventional chain saw, skidder, loader, and tandemaxle truck operations in eastern mountain areas. Production of each function of the logging system may be determined so that the system may be balanced for minimum cost. The...

  10. X-ray table

    International Nuclear Information System (INIS)

    Craig, J.R.; Otto, G.W.

    1980-01-01

    An X-ray radiographic or fluoroscopic table is described which includes a film holder with a frame attached to a cable running over end pulleys for positioning the holder longitudinally as desired under the table top. The holder has a front opening to receive a cassette-supporting tray which can be slid out on tracks to change the cassette. A reed switch on the frame is opened by a permanent magnet on the tray only when the tray is half-way out. When the switch is closed, an electromagnet locks the pulley and the holder in place. The holder is thus automatically locked in place not only during exposure (tray in) but when the tray is out for changing the cassette. To re-position the holder, the operator pulls the tray half-out and, using the tray itself, pushes the holder along the table, the holder being counterbalanced by a weight. (author)

  11. X-ray equipment

    International Nuclear Information System (INIS)

    Redmayne, I.G.B.

    1988-01-01

    The patent concerns a warning and protection system for mobile x-ray equipment used for 'on site' radiography, so that workers in the vicinity of such a working unit can be alerted to its presence. The invention is a local repeater warning system which gives a preliminary warning that energisation of the tubehead is imminent, as well as a switch near the tubehead to abort or inhibit energisation. The latter switch allows personnel caught in the vicinity of the tubehead to prevent energisation. The preliminary warning may be flashing lamps or by a klaxon. The control unit for the equipment may include a monitoring circuit to detect failure of the warning light or klaxon. (U.K.)

  12. X-ray equipment

    Energy Technology Data Exchange (ETDEWEB)

    Redmayne, I.G.B.

    1988-01-06

    The patent concerns a warning and protection system for mobile x-ray equipment used for 'on site' radiography, so that workers in the vicinity of such a working unit can be alerted to its presence. The invention is a local repeater warning system which gives a preliminary warning that energisation of the tubehead is imminent, as well as a switch near the tubehead to abort or inhibit energisation. The latter switch allows personnel caught in the vicinity of the tubehead to prevent energisation. The preliminary warning may be flashing lamps or by a klaxon. The control unit for the equipment may include a monitoring circuit to detect failure of the warning light or klaxon. (U.K.).

  13. X-Ray Lasers 2016

    CERN Document Server

    Bulanov, Sergei; Daido, Hiroyuki; Kato, Yoshiaki

    2018-01-01

    These proceedings comprise a selection of invited and contributed papers presented at the 15th International Conference on X-Ray Lasers (ICXRL 2016), held at the Nara Kasugano International Forum, Japan, from May 22 to 27, 2016. This conference was part of an ongoing series dedicated to recent developments in the science and technology of x-ray lasers and other coherent x-ray sources with additional focus on supporting technologies, instrumentation and applications.   The book showcases recent advances in the generation of intense, coherent x-rays, the development of practical devices and their applications across a wide variety of fields. It also discusses emerging topics such as plasma-based x-ray lasers, 4th generation accelerator-based sources and higher harmonic generations, as well as other x-ray generation schemes.

  14. X-ray instrumentation in astronomy

    International Nuclear Information System (INIS)

    Cuhlane, J.L.

    1985-01-01

    This book presents the proceedings of a conference devoted to x-ray instrumentation in astronomy. Special sections are: AXAF X-Ray Optical Systems; Specialized X-Ray Systems; X-Ray Optical Systems I; X-Ray Optical Systems II; Gas Filled X-Ray Detectors II; The NASA Advanced X-Ray Astrophysics Facility; X-Ray and EUV Spectrometers; Microchannel Plates; and Solid State Detectors

  15. Topological X-Rays Revisited

    Science.gov (United States)

    Lynch, Mark

    2012-01-01

    We continue our study of topological X-rays begun in Lynch ["Topological X-rays and MRI's," iJMEST 33(3) (2002), pp. 389-392]. We modify our definition of a topological magnetic resonance imaging and give an affirmative answer to the question posed there: Can we identify a closed set in a box by defining X-rays to probe the interior and without…

  16. X-ray Absorption Spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Yano, Junko; Yachandra, Vittal K.

    2009-07-09

    This review gives a brief description of the theory and application of X-ray absorption spectroscopy, both X-ray absorption near edge structure (XANES) and extended X-ray absorption fine structure (EXAFS), especially, pertaining to photosynthesis. The advantages and limitations of the methods are discussed. Recent advances in extended EXAFS and polarized EXAFS using oriented membranes and single crystals are explained. Developments in theory in understanding the XANES spectra are described. The application of X-ray absorption spectroscopy to the study of the Mn4Ca cluster in Photosystem II is presented.

  17. X-ray filtration apparatus

    International Nuclear Information System (INIS)

    Thompson, G.

    1992-01-01

    This invention relates to an X-ray shielding support device. In spite of considerable development in X-ray taking techniques, a need still exists for effective shielding, inter alia, to compensate for variations in the thickness, density and the absorption properties of the object being studied. By appropriate shielding, the X-ray image produced is of sufficient detail, contrast and intensity over its entire area to constitute a useful diagnostic aid. It is also desirable to subject the patient to the smallest possible X-ray dosage. 4 figs

  18. X-ray emission spectroscopy. X-ray fluorescence

    International Nuclear Information System (INIS)

    Despujols, J.

    1992-01-01

    Principles of X-ray emission spectrometry are first recalled, then wave-length dispersive and energy dispersive X-ray fluorescence spectrometer are described. They are essentially designed for qualitative and quantitative analysis of elements (Z>10). Sample preparation, calibration, corrections, interferences, accuracy are reviewed. Examples of use in different industries are given. (71 refs.)

  19. Automatic segmentation of mandible in panoramic x-ray

    OpenAIRE

    Abdi, Amir Hossein; Kasaei, Shohreh; Mehdizadeh, Mojdeh

    2015-01-01

    As the panoramic x-ray is the most common extraoral radiography in dentistry, segmentation of its anatomical structures facilitates diagnosis and registration of dental records. This study presents a fast and accurate method for automatic segmentation of mandible in panoramic x-rays. In the proposed four-step algorithm, a superior border is extracted through horizontal integral projections. A modified Canny edge detector accompanied by morphological operators extracts the inferior border of t...

  20. Time switch for X-ray diagnosis apparatus

    Energy Technology Data Exchange (ETDEWEB)

    Distler, G; Goetzl, H

    1977-04-07

    A time switch for dental X-rays consists of a knob to select exposure time. Two scales are concentrically mounted, one for time, and one with the various tooth symbols, in such a way that the various teeth are in correspondence with the usually recommended times. However, should the X-ray tube characteristics vary at some stage, by pressing the knob, then turning, the 'tooth scale' can be rotated for setting the exposure times at higher or lower levels.

  1. The Application of Scanning Electron Microscopy with Energy-Dispersive X-Ray Spectroscopy (SEM-EDX) in Ancient Dental Calculus for the Reconstruction of Human Habits

    Czech Academy of Sciences Publication Activity Database

    Fialová, D.; Skoupý, Radim; Drozdová, E.; Paták, Aleš; Piňos, Jakub; Šín, L.; Beňuš, R.; Klíma, B.

    2017-01-01

    Roč. 23, č. 6 (2017), s. 1207-1213 ISSN 1431-9276 R&D Projects: GA MŠk(CZ) LO1212; GA MŠk ED0017/01/01 Institutional support: RVO:68081731 Keywords : ancient dental calculus * SEM-EDX * human habits * the Great Moravian Empire * Napoleonic Wars Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering OBOR OECD: Electrical and electronic engineering Impact factor: 1.891, year: 2016

  2. 21 CFR 872.1800 - Extraoral source x-ray system.

    Science.gov (United States)

    2010-04-01

    ... (CONTINUED) MEDICAL DEVICES DENTAL DEVICES Diagnostic Devices § 872.1800 Extraoral source x-ray system. (a... dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Extraoral source x-ray system. 872.1800 Section...

  3. Radiation hazards and radiation protection in dental X-ray diagnostics. Guideline for students of dentistry, dental orthopedics and dentists. Strahlengefaehrdung und Strahlenschutz in der zahnaerztlichen Roentgendiagnostik. Leitfaden fuer Studierende der Zah-Mund-Kieferheilkunde und Zahnaerzte

    Energy Technology Data Exchange (ETDEWEB)

    Kirsch, T [ed.

    1977-01-01

    Questions concerning technical-radiological-practical activities with regard to radiation hazards for patient and dentist as well as radiation protection measures are dealt with comprehensively. Recent findings in the field of radiation physics and radiation genetics are taken into account. The annex contains the text of the X-ray Ordinance.

  4. Hand-held transendoscopic robotic manipulators: A transurethral laser prostate surgery case study.

    Science.gov (United States)

    Hendrick, Richard J; Mitchell, Christopher R; Herrell, S Duke; Webster, Robert J

    2015-11-01

    Natural orifice endoscopic surgery can enable incisionless approaches, but a major challenge is the lack of small and dexterous instrumentation. Surgical robots have the potential to meet this need yet often disrupt the clinical workflow. Hand-held robots that combine thin manipulators and endoscopes have the potential to address this by integrating seamlessly into the clinical workflow and enhancing dexterity. As a case study illustrating the potential of this approach, we describe a hand-held robotic system that passes two concentric tube manipulators through a 5 mm port in a rigid endoscope for transurethral laser prostate surgery. This system is intended to catalyze the use of a clinically superior, yet rarely attempted, procedure for benign prostatic hyperplasia. This paper describes system design and experiments to evaluate the surgeon's functional workspace and accuracy using the robot. Phantom and cadaver experiments demonstrate successful completion of the target procedure via prostate lobe resection.

  5. A study of cladding technology on tube wall surface by a hand-held laser torch

    International Nuclear Information System (INIS)

    Terada, Takaya; Nishimura, Akihiko; Oka, Kiyoshi; Moriyama, Taku; Matsuda, Hiroyasu

    2015-01-01

    New maintenance technique was proposed using a hand-held laser torch for aging chemical plants and power plants. The hand-held laser torch was specially designed to be able to access limited tubular space in various cases. A composite-type optical fiberscope was composed of a center fiber for beam delivery and surrounded fibers for visible image delivery. Laser irradiation on a work pieces with the best accuracy of filler wire was carried out. And, we found that the optimized wire-feed speed was 2 mm/s in laser cladding. We succeeded to make a line clad on the inner wall of 23 mm tube. This technique was discussed to be applied to the maintenance for cracks or corrosions of tubes in various harsh environments. (author)

  6. Hand held lasers, a hazard to aircraft: How do we address this?

    Science.gov (United States)

    Barat, K.

    2015-10-01

    The availability of hand held lasers, commonly termed "laser pointers" is easy and wide spread, through commercial web sites and brick & mortar stores. The output of these hand held devices ranges from 1-5 milliWatts (mW) the legal laser pointer output limit, to 5000mW (5Watts). This is thousand times the maximum limit for pointers. Sadly the abuse of these devices is also wide spread. Over the last few years over 3000 aircraft are exposed to laser hits per year. While these aircraft exposures are of no danger to the aircraft frame but they can cause pilot distractions with the potential to cause a serve accident. The presentation will discuss the problem review visual effects, the regulatory response and how educators need to be aware of the problem and can take steps to educate students in the hope of having an effect.

  7. X-ray photoelectron spectroscopy

    International Nuclear Information System (INIS)

    Attekum, P.M.T.M. van.

    1979-01-01

    The methods and results of X-ray photoelectron spectroscopy in the study of plasmons, alloys and gold compounds are discussed. After a comprehensive introduction, seven papers by the author, previously published elsewhere, are reprinted and these cover a wide range of the uses of X-ray photoelectron spectroscopy. (W.D.L.)

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... over time. top of page What are the benefits vs. risks? Benefits Bone x-rays are the fastest and easiest ... bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams ...

  9. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that those parts of a patient's body not being imaged receive minimal radiation exposure. top of page What are the limitations of Bone X-ray (Radiography)? ...

  10. Traditional x-ray imaging

    International Nuclear Information System (INIS)

    Hay, G.A.

    1982-01-01

    Methods of imaging x-rays, with particular reference to medicine, are reviewed. The history and nature of x-rays, their production and spectra, contrast, shapes and fine structure, image transducers, including fluorescent screens, radiography, fluoroscopy, and image intensifiers, image detection, perception and enhancement and clinical applications are considered. (U.K.)

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that those parts of a patient's body not being imaged receive minimal radiation exposure. ...

  12. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... have special pediatric considerations. The teddy bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams Arthritis X-ray, Interventional Radiology and Nuclear Medicine Radiation Safety How to Read Your Radiology Report ...

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... are a form of radiation like light or radio waves. X-rays pass through most objects, including the body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small burst of radiation that passes through the body, recording ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small burst of radiation that passes through the body, recording an image on photographic film or a special detector. Different parts of the body absorb the x-rays in ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... clothing that might interfere with the x-ray images. Women should always inform their physician and x-ray ... lowest radiation dose possible while producing the best images for ... organizations continually review and update the technique standards used ...

  16. Changes in bananas and guavas after irradiation by dental X-ray equipment; Alterações em bananas e goiabas após irradiação por equipamento odontológico de raios X

    Energy Technology Data Exchange (ETDEWEB)

    Santos, S.M.; Silva, R.S., E-mail: simonemattosantos@gmail.com [Universidade do Grande Rio (UNIGRANRIO), Duque de Caxias, RJ (Brazil). Curso Superior de Tecnologia em Radiologia; Oliveira, J.S.; Gomes, A.S [Fculdade Casa Branca (FACAB), SP (Brazil). Pós-graduação em Proteção Radiológica em Aplicações Médicas, Industriais e Nucleares

    2017-07-01

    Food irradiation is a process in which foods are subjected to controlled doses of ionizing radiation. The main advantage is the destruction of microorganisms, maintaining their characteristics longer. The objective of this study was to verify if the low energy range of X - rays emitted by periapical dental equipment is capable of generating some alteration in the maturation and / or decay time of fruits such as bananas and guavas. The methodology included the purchase of bananas and green guavas, transported to the Laboratory of Dental Radiology of the University of Grande Rio, to expose the existing equipment in the facility. A batch of bananas was separated to be irradiated and one not to be irradiated (control group). The same happened with two guavas. Each type of fruit was irradiated on 06/18/2016 at 2:35 pm, selecting 70 kVp and 15 mAs on the control panel. Fruit characteristics, such as color and texture, were observed and recorded for 17 days. It was observed on the 14{sup th} day that the bananas of the control group presented more yellowish color and brown color, indicating a high degree of ripening and at the end of the stipulated time they showed rot. The irradiated bananas presented brown coloration, but without signs of rotting. Guavas, irradiated and control, were rotted at the end of the 17 days, but control at a higher grade. It is concluded that even the low energetic band of the beam emitted by the periapical dental equipment is able to alter the time of maturation and decay of the fruits. Non-utility for commercial purposes is emphasized.

  17. Precise intraoperative location of gastrointestinal bleeding with a hand-held counter. Work in progress

    International Nuclear Information System (INIS)

    Williamson, M.R.; Boyd, C.M.; McGuire, E.L.

    1986-01-01

    The nuclear medicine bleeding scan is frequently insufficient to locate sites of bleeding precisely, in spite of its great sensitivity. A small, hand-held Geiger-Mueller counter, placed directly on exposed intestine in the operating room, enables precise location of the probable bleeding site. In three patients, the technique allowed a minimal amount of intestine to be resected, distinguished between large- and small-intestinal hemorrhage, and eliminated other foci as sites of bleeding.A

  18. X-ray imaging system

    International Nuclear Information System (INIS)

    Houston, J.M.

    1980-01-01

    A novel, high-speed apparatus for use in X-ray computerised tomography is described in detail. It consists of a semi-circular array of X-ray sources, collimators and an ion chamber array for detection of the X-rays. The X-ray sources may be pulsed in salvos such that the corresponding detectors in the array are only illuminated by one source. The use of computer controlled salvos speeds up the image processing by at least a factor of two. The ion chamber array is designed to have a constant detection efficiency for varying angles of X-ray incidence. A detailed description of the detector construction and suggested gaseous fillings are given. It is claimed that the present tomographic system allows fast and accurate imaging of internal body organs and is insensitive to the blurring effects which motion of these organs tends to produce. (UK)

  19. Validity of maximal isometric knee extension strength measurements obtained via belt-stabilized hand-held dynamometry in healthy adults.

    Science.gov (United States)

    Ushiyama, Naoko; Kurobe, Yasushi; Momose, Kimito

    2017-11-01

    [Purpose] To determine the validity of knee extension muscle strength measurements using belt-stabilized hand-held dynamometry with and without body stabilization compared with the gold standard isokinetic dynamometry in healthy adults. [Subjects and Methods] Twenty-nine healthy adults (mean age, 21.3 years) were included. Study parameters involved right side measurements of maximal isometric knee extension strength obtained using belt-stabilized hand-held dynamometry with and without body stabilization and the gold standard. Measurements were performed in all subjects. [Results] A moderate correlation and fixed bias were found between measurements obtained using belt-stabilized hand-held dynamometry with body stabilization and the gold standard. No significant correlation and proportional bias were found between measurements obtained using belt-stabilized hand-held dynamometry without body stabilization and the gold standard. The strength identified using belt-stabilized hand-held dynamometry with body stabilization may not be commensurate with the maximum strength individuals can generate; however, it reflects such strength. In contrast, the strength identified using belt-stabilized hand-held dynamometry without body stabilization does not reflect the maximum strength. Therefore, a chair should be used to stabilize the body when performing measurements of maximal isometric knee extension strength using belt-stabilized hand-held dynamometry in healthy adults. [Conclusion] Belt-stabilized hand-held dynamometry with body stabilization is more convenient than the gold standard in clinical settings.

  20. The reliability and validity of hand-held refractometry water content measures of hydrogel lenses.

    Science.gov (United States)

    Nichols, Jason J; Mitchell, G Lynn; Good, Gregory W

    2003-06-01

    To investigate within- and between-examiner reliability and validity of hand-held refractometry water content measures of hydrogel lenses. Nineteen lenses of various nominal water contents were examined by two examiners on two occasions separated by 1 hour. An Atago N2 hand-held refractometer was used for all water content measures. Lenses were presented in a random order to each examiner by a third party, and examiners were masked to any potential lens identifiers. Intraclass correlation coefficients (ICC), 95% limits of agreement, and Wilcoxon signed rank test were used to characterize the within- and between-examiner reliability and validity of lens water content measures. Within-examiner reliability was excellent (ICC, 0.97; 95% limits of agreement, -3.6% to +5.7%), and the inter-visit mean difference of 1.1 +/- 2.4% was not biased (p = 0.08). Between-examiner reliability was also excellent (ICC, 0.98; 95% limits of agreement, -4.1% to +3.9%). The mean difference between examiners was -0.1 +/- 2.1% (p = 0.83). The mean difference between the nominally reported water content and our water content measures was -2.1 +/- 1.7% (p refractometry and is material dependent. Therefore, investigators may need to account for bias when measuring hydrogel lens water content via hand-held refractometry.

  1. New portable hand-held radiation instruments for measurements and monitoring

    International Nuclear Information System (INIS)

    Fehlau, P.E.

    1987-01-01

    Hand-held radiation monitors are often used to search pedestrians and motor vehicles for special nuclear material (SNM) as part of a physical protection plan for nuclear materials. Recently, the Los Alamos Advanced Nuclear Technology group has commercialized an improved hand-held monitor that can be used for both physical-protection monitoring and verification measurements in nuclear material control and waste management. The new monitoring instruments are smaller and lighter; operate much longer on a battery charge; are available with NaI(Tl) or neutron and gamma-ray sensitive plastic scintillation detectors; and are less expensive than other comparable instruments. They also have a second operating mode for making precise measurements over counting times as long as 99 s. This mode permits making basic verification measurements that may be needed before transporting nuclear material or waste outside protected areas. Improved verification measurements can be made with a second new hand-held instrument that has a stabilized detector and three separate gamma-ray energy windows to obtain spectral information for SNM quantity, enrichment, or material-type verification

  2. X-ray diagnostics for TFTR

    International Nuclear Information System (INIS)

    von Goeler, S.; Hill, K.W.; Bitter, M.

    1982-12-01

    A short description of the x-ray diagnostic preparation for the TFTR tokamak is given. The x-ray equipment consists of the limiter x-ray monitoring system, the soft x-ray pulse-height-analysis-system, the soft x-ray imaging system and the x-ray crystal spectrometer. Particular attention is given to the radiation protection of the x-ray systems from the neutron environment

  3. X-ray filter for chest X-rays

    International Nuclear Information System (INIS)

    Ferlic, D.J.

    1984-01-01

    A description is given of an X-ray filter comprised of a sheet of radiation absorbing material with an opening corresponding to the spine and central portion of the heart. The upper portion of the filter exhibits a relatively narrow opening which becomes gradually wider toward the lower portion of the filter. This filter will permit an acceptable density level of x-ray exposure for the lungs while allowing a higher level of x-ray exposure for the mediastinum areas of the body. (author)

  4. X-ray filter for chest x-rays

    International Nuclear Information System (INIS)

    Ferlic, D.J.

    1984-01-01

    Filter for use in medical x-ray apparatus to permit higher intensity x-ray exposure in the heart and mediastinum area while maintaining a normal level of x-ray exposure in other areas of the body, particlarly in the lung area. The filter comprises a sheet of radiation absorbing material having an opening therein, said opening corresponding to the spine and central portion of the heart. Accordingly, the upper portion of the filter exhibits a relatively narrow opening which becomes gradually wider toward the lower portion of the filter

  5. Semiconductor X-ray detectors

    CERN Document Server

    Lowe, Barrie Glyn

    2014-01-01

    Identifying and measuring the elemental x-rays released when materials are examined with particles (electrons, protons, alpha particles, etc.) or photons (x-rays and gamma rays) is still considered to be the primary analytical technique for routine and non-destructive materials analysis. The Lithium Drifted Silicon (Si(Li)) X-Ray Detector, with its good resolution and peak to background, pioneered this type of analysis on electron microscopes, x-ray fluorescence instruments, and radioactive source- and accelerator-based excitation systems. Although rapid progress in Silicon Drift Detectors (SDDs), Charge Coupled Devices (CCDs), and Compound Semiconductor Detectors, including renewed interest in alternative materials such as CdZnTe and diamond, has made the Si(Li) X-Ray Detector nearly obsolete, the device serves as a useful benchmark and still is used in special instances where its large, sensitive depth is essential. Semiconductor X-Ray Detectors focuses on the history and development of Si(Li) X-Ray Detect...

  6. X-ray film calibration

    International Nuclear Information System (INIS)

    Stone, G.F.; Dittmore, C.H.; Henke, B.L.

    1986-01-01

    This paper discusses the use of silver halide x-ray films for imaging and spectroscopy which is limited by the range of intensities that can be recorded and densitometered. Using the manufacturers processing techniques can result in 10 2-3 range in intensity recorded over 0-5 density range. By modifying the chemistry and processing times, ranges of 10 5-6 can be recorded in the same density range. The authors report on x-ray film calibration work and dynamic range improvements. Changes to the processing chemistry and the resulting changes in dynamic range and x-ray sensitivity are discussed

  7. Women and x-rays

    Energy Technology Data Exchange (ETDEWEB)

    Dunkley, P A; Stewart, J H

    1976-01-01

    When a woman comes to an X-Ray Department it is usually necessary to know the present stage of her menstrual cycle. X-Rays may have an adverse effect on the embryo, especially in early pregnancy. However, exposure to X-Rays at any stage may be associated with a slightly increased incidence of malignant disease in childhood. The International Commission on Radiological Protection recommends that in women of child-bearing age (in some cases as young as 11 years), non-urgent diagnostic radiography be confined to the preovulatory phase of the menstrual cycle: that is, 14 days following the first day of the last menstrual period.

  8. Upsurge of X-ray astronomy 230-

    International Nuclear Information System (INIS)

    Hudec, D.R.

    1978-01-01

    Instruments are described used for X-ray astronomy, namely X-ray detectors and X-ray telescopes. Unlike telescopes, the detectors do not comprise X-ray optics. A survey is given of the results obtained in solar and stellar X-ray astronomy and hypotheses are submitted on the origin of X radiation in the interstellar space. (J.B.)

  9. A compliance testing program for diagnostic X-ray equipment

    International Nuclear Information System (INIS)

    Hutchinson, D.E.; Cobb, B.J.; Jacob, C.S.

    1999-01-01

    Compliance testing is nominally that part of a quality assurance program dealing with those aspects of X-ray equipment performance that are subject to radiation control legislation. Quality assurance programs for medical X-ray equipment should be an integral part of the quality culture in health care. However while major hospitals and individual medical centers may implement such programs with some diligence, much X-ray equipment can remain unappraised unless there is a comprehensive regulatory inspection program or some form of compulsion on the equipment owner to implement a testing program. Since the late 1950s all X-ray equipment in the State of Western Australia has been inspected by authorized officers acting on behalf of the Radiological Council, the regulatory authority responsible for administration of the State's Radiation Safety Act. However, economic constraints, coupled with increasing X-ray equipment numbers and a geographically large State have significantly affected the inspection rate. Data available from inspections demonstrate that regular compliance and performance checks are essential in order to ensure proper performance and to minimize unnecessary patient and operator dose. To ensure that diagnostic X-ray equipment complies with accepted standards and performance criteria, the regulatory authority introduced a compulsory compliance testing program for all medical, dental and chiropractic diagnostic X-ray equipment effective from 1 January 1997

  10. Evaluation of crystalline changes and resistance to demineralization of the surface of human dental enamel treated with Er:YAG laser and fluoride using x-ray diffraction analysis and Vickers microhardness

    Science.gov (United States)

    Behroozibakhsh, Marjan; Shahabi, Sima; Ghavami-Lahiji, Mehrsima; Sadeghian, Safura; Sadat Faal Nazari, Neda

    2018-06-01

    This study aimed to investigate the changes in crystalline structure and resistance to demineralization of human dental surface enamel treated with erbium-doped yttrium aluminium garnet laser (Er:YAG) laser and fluoride. The enamel surfaces were divided into four groups according to the treatment process including, (L): irradiated with Er:YAG; (F): treated with acidulated phosphate fluoride gel (LF): Pre-irradiated surfaces with Er:YAG subjected to acidulated phosphate fluoride gel and (FL): laser irradiation was performed on the fluoridated enamel surface. Before and after the treatment procedure, the samples were evaluated using X-ray diffraction, scanning electron microscope (SEM) and the Vickers microhardness test. The surface microhardness values also were measured after a pH-cycling regime and acid challenge. The a-axis of all lased groups was contracted after treatment procedure. Measurement of the area under the peaks showed the highest crysallinity in the FL group. The hardness values of all laser treated samples significantly reduced after treatment procedure compared to the F group (p  ⩽  0.001). The morphological observations showed remarkable changes on the lased enamel surfaces including cracks, craters and exposed prisms. These findings suggest, irradiation of the Er:YAG laser accompanying with fluoride application can induce some beneficial crystalline changes regarding the acid-resistance properties of enamel, however, the craters and cracks produced by laser irradiation can promote enamel demineralization and consequently the positive effects of the Er:YAG laser will be eliminated.

  11. X-ray energy selected imaging with Medipix II

    International Nuclear Information System (INIS)

    Ludwig, J.; Zwerger, A.; Benz, K.-W.; Fiederle, M.; Braml, H.; Fauler, A.; Konrath, J.-P.

    2004-01-01

    Two different X-ray tube accelerating voltages (60 and 70 kV) are used for diagnosis of front teeth and molars. Different energy ranges are necessary as function of tooth thickness to obtain similar contrast for imaging. This technique drives the costs for the X-ray tube up and allows for just two optimized settings. Energy range selection for the detection of the penetrating X-rays would overcome these severe setbacks. The single photon counting chip MEDIPIX2 http://www.cern.ch/medipix exhibits exactly this feature. First simulations and measurements have been carried out using a dental X-ray source. As a demonstrator a real tooth has been used with different cavities and filling materials. Simulations showed in general larger improvements as compared to measurements regarding SNR and contrast: A beneficial factor of 4% wrt SNR and 25% for contrast, measurements showed factors of 2.5 and up to 10%, respectively

  12. X-ray energy selected imaging with Medipix II

    Science.gov (United States)

    Ludwig, J.; Zwerger, A.; Benz, K.-W.; Fiederle, M.; Braml, H.; Fauler, A.; Konrath, J.-P.

    2004-09-01

    Two different X-ray tube accelerating voltages (60 and 70kV) are used for diagnosis of front teeth and molars. Different energy ranges are necessary as function of tooth thickness to obtain similar contrast for imaging. This technique drives the costs for the X-ray tube up and allows for just two optimized settings. Energy range selection for the detection of the penetrating X-rays would overcome these severe setbacks. The single photon counting chip MEDIPIX2 http://www.cern.ch/medipix exhibits exactly this feature.First simulations and measurements have been carried out using a dental X-ray source. As a demonstrator a real tooth has been used with different cavities and filling materials. Simulations showed in general larger improvements as compared to measurements regarding SNR and contrast: A beneficial factor of 4% wrt SNR and 25% for contrast, measurements showed factors of 2.5 and up to 10%, respectively.

  13. The IHS diagnostic X-ray equipment radiation protection program

    International Nuclear Information System (INIS)

    Knapp, A.; Byrns, G.; Suleiman, O.

    1994-01-01

    The Indian Health Service (IHS) operates or contracts with Tribal groups to operate 50 hospitals and approximately 165 primary ambulatory care centers. These facilities contain approximately 275 medical and 800 dental diagnostic x-ray machines. IHS environmental health personnel in collaboration with the Food and Drug Administration's (FDA) Center for Devices and Radiological Health (CDRH) developed a diagnostic x-ray protection program including standard survey procedures and menu-driven calculations software. Important features of the program include the evaluation of equipment performance collection of average patient entrance skin exposure (ESE) measurements for selected procedures, and quality assurance. The ESE data, collected using the National Evaluation of X-ray Trends (NEXT) protocol, will be presented. The IHS Diagnostic X-ray Radiation Protection Program is dynamic and is adapting to changes in technology and workload

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... more information about pregnancy and x-rays. A Word About Minimizing Radiation Exposure Special care is taken ... and/or your insurance provider to get a better understanding of the possible charges you will incur. ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... is repeated. Two or three images (from different angles) will typically be taken. An x-ray may ... RadiologyInfo.org is not a medical facility. Please contact your physician with specific medical questions or for ...

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... no special preparation. Tell your doctor and the technologist if there is any possibility you are pregnant. ... should always inform their physician and x-ray technologist if there is any possibility that they are ...

  17. X-ray guided biopsy

    International Nuclear Information System (INIS)

    Casanova, R.; Lezana, A.H.; Pedrosa, C.S.

    1980-01-01

    Fine needle aspiration biopsy (FNAB) is now a routine procedure in many X-ray Departments. This paper presents the authors' experience with this technique in chest, abdominal and skeletal lesions. (Auth.)

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... Radiography) - Bone Bone x-ray uses a very small dose of ionizing radiation to produce pictures of ... exposing a part of the body to a small dose of ionizing radiation to produce pictures of ...

  19. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... replacement and fracture reductions. look for injury, infection, arthritis , abnormal bone growths and bony changes seen in ... injuries, including fractures, and joint abnormalities, such as arthritis. X-ray equipment is relatively inexpensive and widely ...

  20. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... procedure varies. See the Safety page for more information about radiation dose. Women should always inform their ...

  1. X-Ray Assembler Data

    Data.gov (United States)

    U.S. Department of Health & Human Services — Federal regulations require that an assembler who installs one or more certified components of a diagnostic x-ray system submit a report of assembly. This database...

  2. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... technologist, an individual specially trained to perform radiology examinations, positions the patient on the x-ray table ... bone is forming), for comparison purposes. When the examination is complete, you may be asked to wait ...

  3. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... in a known abnormality can be monitored over time. Follow-up examinations are sometimes the best way ...

  4. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... the radiation while soft tissue, such as muscle, fat and organs, allow more of the x-rays ... taken of the unaffected limb, or of a child's growth plate (where new bone is forming), for ...

  5. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... the radiation while soft tissue, such as muscle, fat and organs, allow more of the x-rays ... information you were looking for? Yes No Please type your comment or suggestion into the following text ...

  6. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... pregnant. Many imaging tests are not performed during pregnancy so as not to expose the fetus to ... See the Safety page for more information about pregnancy and x-rays. top of page What does ...

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... a large photographic negative). Today, most images are digital files that are stored electronically. These stored images ... and places the x-ray film holder or digital recording plate under the table in the area ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... in metabolic conditions. assist in the detection and diagnosis of bone cancer . locate foreign objects in soft ... frequently compared to current x-ray images for diagnosis and disease management. top of page How is ...

  9. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... of any bone in the body, including the hand, wrist, arm, elbow, shoulder, spine, pelvis, hip, thigh, knee, leg ( ... Image Gallery Radiological technologist preparing to take an arm x-ray on a ... Images related ...

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... care is taken during x-ray examinations to use the lowest radiation dose possible while producing the best images for evaluation. National and international radiology protection organizations continually review ...

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that ... radiation oncology provider in your community, you can search the ACR-accredited facilities database . This website does ...

  12. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... small burst of radiation that passes through the body, recording an image on photographic film or a special detector. Different parts of the body absorb the x-rays in varying degrees. Dense ...

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... x-ray machine is a compact apparatus that can be taken to the patient in a hospital ... so that any change in a known abnormality can be monitored over time. Follow-up examinations are ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... of page What will I experience during and after the procedure? A bone x-ray examination itself ... available in emergency rooms, physician offices, ambulatory care centers, nursing homes and other locations, making it convenient ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... will analyze the images and send a signed report to your primary care or referring physician , who ... Medicine Radiation Safety How to Read Your Radiology Report Images related to X-ray (Radiography) - Bone Sponsored ...

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams ... the web pages found at these links. About Us | Contact Us | FAQ | Privacy | Terms of Use | Links | ...

  17. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... to produce pictures of any bone in the body. It is commonly used to diagnose fractured bones ... x-rays involves exposing a part of the body to a small dose of ionizing radiation to ...

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... to X-ray (Radiography) - Bone Sponsored by Please note RadiologyInfo.org is not a medical facility. Please ... is further reviewed by committees from the American College of Radiology (ACR) and the Radiological Society of ...

  19. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... the baby. See the Safety page for more information about pregnancy and x-rays. top of page ... procedure varies. See the Safety page for more information about radiation dose. Women should always inform their ...

  20. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... and fracture reductions. look for injury, infection, arthritis , abnormal bone growths and bony changes seen in metabolic ... to current x-ray images for diagnosis and disease management. top of page How is the procedure ...

  1. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... imaged. When necessary, sandbags, pillows or other positioning devices will be used to help you maintain the ... here Images × Image Gallery Radiological technologist preparing to take an arm x-ray on a patient. View ...

  2. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... foot. top of page What are some common uses of the procedure? A bone x-ray is ... community, you can search the ACR-accredited facilities database . This website does not provide cost information. The ...

  3. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... patient. top of page How does the procedure work? X-rays are a form of radiation like ... taken of the unaffected limb, or of a child's growth plate (where new bone is forming), for ...

  4. Flash x-ray cinematography

    International Nuclear Information System (INIS)

    Stein, W.E.

    1976-01-01

    Experiments intended to provide an overview of the potential capabilities and limitations of flash x-ray cinematography as a diagnostic technique for a Fast Reactor Safety Test Facility are described. The results provide estimates of the x-ray pulse intensity required to obtain adequate radiographs of an array of fuel pins in a typical reactor configuration. An estimate of the upper limit on the pulse duration imposed by the reactor background radiation was also determined. X-ray cinematography has been demonstrated at a repetition rate limited only by the recording equipment on hand at the time of these measurements. These preliminary results indicate that flash x-ray cinematography of the motion of fuel in a Fast Reactor Test Facility is technically feasible

  5. X-ray screening materials

    International Nuclear Information System (INIS)

    Wardley, R.B.

    1981-01-01

    This invention relates to x-ray screening materials and especially to materials in sheet form for use in the production of, for example, protective clothing such as aprons and lower back shields, curtains, mobile screens and suspended shields. The invention is based on the observation that x-ray screening materials in sheet form having greater flexiblity than the hitherto known x-ray screening materials of the same x-ray absorber content can be produced if, instead of using a single sheet of filled sheet material of increased thickness, one uses a plurality of sheets of lesser thickness together forming a laminar material of the desired thickness and one bonds the individual sheets together at their edges and, optionally, at other spaced apart points away from the edges thereby allowing one sheet to move relative to another. (U.K.)

  6. X-ray luminescent glasses

    International Nuclear Information System (INIS)

    Takahashi, T.; Yamada, O.

    1981-01-01

    X-ray luminescent glasses comprising a divalent cation such as an alkaline earth metal or other divalent cations such as pb, cd, or zn, and certain rare earth metaphosphates are suitable as vitreous, x-ray phosphors or x-ray luminescent glass fibers in an x-ray intensifying screen. The glasses have the composition n(Mo X p2o5)((1-y)tb2o3 X yce2o3 X 3p2o5) wherein N is greater than zero but less than or equal to 16, M is an alkaline earth metal or other divalent cation such as pb, cd, or zn, and Y is greater than or equal to zero but less than one

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... x-ray uses a very small dose of ionizing radiation to produce pictures of any bone in the ... of the body to a small dose of ionizing radiation to produce pictures of the inside of the ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... examination may also be necessary so that any change in a known abnormality can be monitored over ...

  9. Miniature x-ray source

    Science.gov (United States)

    Trebes, James E.; Bell, Perry M.; Robinson, Ronald B.

    2000-01-01

    A miniature x-ray source utilizing a hot filament cathode. The source has a millimeter scale size and is capable of producing broad spectrum x-ray emission over a wide range of x-ray energies. The miniature source consists of a compact vacuum tube assembly containing the hot filament cathode, an anode, a high voltage feedthru for delivering high voltage to the cathode, a getter for maintaining high vacuum, a connector for initial vacuum pump down and crimp-off, and a high voltage connection for attaching a compact high voltage cable to the high voltage feedthru. At least a portion of the vacuum tube wall is fabricated from highly x-ray transparent materials, such as sapphire, diamond, or boron nitride.

  10. Duodenal X-ray diagnostics

    International Nuclear Information System (INIS)

    Scheppach, W.

    1982-01-01

    The publication provides an overview of duodenal X-ray diagnostics with the aid of barium meals in 1362 patients. The introducing paragraphs deal with the topographic anatomy of the region and the methodics of X-ray investigation. The chapter entitled ''processes at the duodenum itself'' describes mainly ulcers, diverticula, congenital anomalies, tumors and inflammations. The neighbourhood processes comprise in the first place diseases having their origin at the pancreas and bile ducts. As a conclusion, endoscopic rectograde cholangio-pancreaticography and percutaneous transhepatic cholangiography are pointed out as advanced X-ray investigation methods. In the annex of X-ray images some of the described phenomena are shown in exemplary manner. (orig./MG) [de

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... position possible that still ensures x-ray image quality. top of page Who interprets the results and ... accredited facilities database . This website does not provide cost information. The costs for specific medical imaging tests, ...

  12. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... top of page What are the benefits vs. risks? Benefits Bone x-rays are the fastest and ... in the typical diagnostic range for this exam. Risks There is always a slight chance of cancer ...

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... fracture. guide orthopedic surgery, such as spine repair/fusion, joint replacement and fracture reductions. look for injury, ... CT Exams Arthritis X-ray, Interventional Radiology and Nuclear Medicine Radiation Safety How to Read Your Radiology ...

  14. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... the oldest and most frequently used form of medical imaging. A bone x-ray makes images of any ... a radiologist or other physician. To locate a medical imaging or radiation oncology provider in your community, you ...

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... carefully aimed at the part of the body being examined, an x-ray machine produces a small ... the table in the area of the body being imaged. When necessary, sandbags, pillows or other positioning ...

  16. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... ray examination. X-rays usually have no side effects in the typical diagnostic range for this exam. ... minimize stray (scatter) radiation. This ensures that those parts of a patient's body not being imaged receive ...

  17. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... Leave jewelry at home and wear loose, comfortable clothing. You may be asked to wear a gown. ... appliances, eye glasses and any metal objects or clothing that might interfere with the x-ray images. ...

  18. Magnetic x-ray microdiffraction

    Energy Technology Data Exchange (ETDEWEB)

    Evans, Paul G [Computer-Aided Engineering Center, University of Wisconsin, Madison, WI 53706 (United States); Isaacs, Eric D [Center for Nanoscale Materials, Argonne National Laboratory, Argonne, IL 60439 (United States)

    2006-08-07

    Magnetic x-ray microdiffraction uses the structural specificity of x-ray diffraction to probe complex magnetic structures at the length scales relevant to physical phenomena including domain dynamics and phase transitions. Conventional magnetic crystallography techniques such as neutron or x-ray diffraction lack this spatial resolution. The combination of both reciprocal space and real space resolution with a rich magnetic cross section allows new microscopy techniques to be developed and applied to magnetism at the scale of single domains. Potential applications include a wide range of magnetic problems in nanomagnetism, the interaction of strain, polarization and magnetization in complex oxides and spatially resolved studies of magnetic phase transitions. We present the physical basis for x-ray microdiffraction and magnetic scattering processes, review microdiffraction domain imaging techniques in antiferromagnetic and ferromagnetic materials and discuss potential directions for studies. (topical review)

  19. Electromechanical x-ray generator

    Science.gov (United States)

    Watson, Scott A; Platts, David; Sorensen, Eric B

    2016-05-03

    An electro-mechanical x-ray generator configured to obtain high-energy operation with favorable energy-weight scaling. The electro-mechanical x-ray generator may include a pair of capacitor plates. The capacitor plates may be charged to a predefined voltage and may be separated to generate higher voltages on the order of hundreds of kV in the AK gap. The high voltage may be generated in a vacuum tube.

  20. Accelerator x-ray sources

    CERN Document Server

    Talman, Richard

    2007-01-01

    This first book to cover in-depth the generation of x-rays in particle accelerators focuses on electron beams produced by means of the novel Energy Recovery Linac (ERL) technology. The resulting highly brilliant x-rays are at the centre of this monograph, which continues where other books on the market stop. Written primarily for general, high energy and radiation physicists, the systematic treatment adopted by the work makes it equally suitable as an advanced textbook for young researchers.

  1. X-ray tube target

    International Nuclear Information System (INIS)

    Weber, R.G.

    1980-01-01

    A target with an improved heat emissive surface for use in a rotating anode type x-ray tube is described. The target consists of a body having a first surface portion made of x-ray emissive material and a second surface portion made of a heat emissive material comprising at least one of hafnium boride, hafnium oxide, hafnium nitride, hafnium silicide, and hafnium aluminide. (U.K.)

  2. X-ray fluorescence holography

    CERN Document Server

    Hayashi, K; Takahashi, Y

    2003-01-01

    X-ray fluorescence holography (XFH) is a new structural analysis method of determining a 3D atomic arrangement around fluorescing atoms. We developed an XFH apparatus using advanced X-ray techniques and succeeded in obtaining high-quality hologram data. Furthermore, we introduced applications to the structural analysis of a thin film and the environment around dopants and, discussed the quantitative analysis of local lattice distortion. (author)

  3. X-ray data processing

    OpenAIRE

    Powell, Harold R.

    2017-01-01

    The method of molecular structure determination by X-ray crystallography is a little over a century old. The history is described briefly, along with developments in X-ray sources and detectors. The fundamental processes involved in measuring diffraction patterns on area detectors, i.e. autoindexing, refining crystal and detector parameters, integrating the reflections themselves and putting the resultant measurements on to a common scale are discussed, with particular reference to the most c...

  4. Low energy (soft) x rays

    International Nuclear Information System (INIS)

    Hoshi, Masaharu; Antoku, Shigetoshi; Russell, W.J.; Miller, R.C.; Nakamura, Nori; Mizuno, Masayoshi; Nishio, Shoji.

    1987-05-01

    Dosimetry of low-energy (soft) X rays produced by the SOFTEX Model CMBW-2 was performed using Nuclear Associates Type 30 - 330 PTW, Exradin Type A2, and Shonka-Wyckoff ionization chambers with a Keithley Model 602 electrometer. Thermoluminescent (BeO chip) dosimeters were used with a Harshaw Detector 2000-A and Picoammeter-B readout system. Beam quality measurements were made using aluminum absorbers; exposure rates were assessed by the current of the X-ray tube and by exposure times. Dose distributions were established, and the average factors for non-uniformity were calculated. The means of obtaining accurate absorbed and exposed doses using these methods are discussed. Survival of V79 cells was assessed by irradiating them with soft X rays, 200 kVp X rays, and 60 Co gamma rays. The relative biological effectiveness (RBE) values for soft X rays with 0, 0.2, 0.7 mm added thicknesses of aluminum were 1.6, which were compared to 60 Co. The RBE of 200 kVp X rays relative to 60 Co was 1.3. Results of this study are available for reference in future RERF studies of cell survival. (author)

  5. Potential role of a new hand-held miniature gamma camera in performing minimally invasive parathyroidectomy

    Energy Technology Data Exchange (ETDEWEB)

    Ortega, Joaquin; Lledo, Salvador [University of Valencia, Clinic University Hospital, Department of Surgery, Valencia (Spain); Ferrer-Rebolleda, Jose [Clinic University Hospital, Department of Nuclear Medicine, Valencia (Spain); Cassinello, Norberto [Clinic University Hospital, Unit of Endocrinologic and Bariatric Surgery, Valencia (Spain)

    2007-02-15

    Sestamibi scans have increased the use of minimally invasive parathyroidectomy (MIP) to treat primary hyperparathyroidism (PHPT) when caused by a parathyroid single adenoma. The greatest concern for surgeons remains the proper identification of pathological glands in a limited surgical field. We have studied the usefulness of a new hand-held miniature gamma camera (MGC) when used intraoperatively to locate parathyroid adenomas. To our knowledge this is the first report published on this subject in the scientific literature. Five patients with PHPT secondary to a single adenoma, positively diagnosed by preoperative sestamibi scans, underwent a MIP. A gamma probe for radioguided surgery and the new hand-held MGC were used consecutively to locate the pathological glands. This new MGC has a module composed of a high-resolution interchangeable collimator and a CsI(Na) scintillating crystal. It has dimensions of around 15 cm x 8 cm x 9 cm and weighs 1 kg. The intraoperative assay of PTH (ioPTH) was used to confirm the complete resection of pathological tissue. All cases were operated on successfully by a MIP. The ioPTH confirmed the excision of all pathological tissues. The MGC proved its usefulness in all patients, even in a difficult case in which the first attempt with the gamma probe failed. In all cases it offered real-time accurate intraoperative images. The hand-held MGC is a useful instrument in MIP for PHPT. It may be used to complement the standard tools used to date, or may even replace them, at least in selected cases of single adenomas. (orig.)

  6. Potential role of a new hand-held miniature gamma camera in performing minimally invasive parathyroidectomy

    International Nuclear Information System (INIS)

    Ortega, Joaquin; Lledo, Salvador; Ferrer-Rebolleda, Jose; Cassinello, Norberto

    2007-01-01

    Sestamibi scans have increased the use of minimally invasive parathyroidectomy (MIP) to treat primary hyperparathyroidism (PHPT) when caused by a parathyroid single adenoma. The greatest concern for surgeons remains the proper identification of pathological glands in a limited surgical field. We have studied the usefulness of a new hand-held miniature gamma camera (MGC) when used intraoperatively to locate parathyroid adenomas. To our knowledge this is the first report published on this subject in the scientific literature. Five patients with PHPT secondary to a single adenoma, positively diagnosed by preoperative sestamibi scans, underwent a MIP. A gamma probe for radioguided surgery and the new hand-held MGC were used consecutively to locate the pathological glands. This new MGC has a module composed of a high-resolution interchangeable collimator and a CsI(Na) scintillating crystal. It has dimensions of around 15 cm x 8 cm x 9 cm and weighs 1 kg. The intraoperative assay of PTH (ioPTH) was used to confirm the complete resection of pathological tissue. All cases were operated on successfully by a MIP. The ioPTH confirmed the excision of all pathological tissues. The MGC proved its usefulness in all patients, even in a difficult case in which the first attempt with the gamma probe failed. In all cases it offered real-time accurate intraoperative images. The hand-held MGC is a useful instrument in MIP for PHPT. It may be used to complement the standard tools used to date, or may even replace them, at least in selected cases of single adenomas. (orig.)

  7. Comparative Geometrical Accuracy Investigations of Hand-Held 3d Scanning Systems - AN Update

    Science.gov (United States)

    Kersten, T. P.; Lindstaedt, M.; Starosta, D.

    2018-05-01

    Hand-held 3D scanning systems are increasingly available on the market from several system manufacturers. These systems are deployed for 3D recording of objects with different size in diverse applications, such as industrial reverse engineering, and documentation of museum exhibits etc. Typical measurement distances range from 0.5 m to 4.5 m. Although they are often easy-to-use, the geometric performance of these systems, especially the precision and accuracy, are not well known to many users. First geometrical investigations of a variety of diverse hand-held 3D scanning systems were already carried out by the Photogrammetry & Laser Scanning Lab of the HafenCity University Hamburg (HCU Hamburg) in cooperation with two other universities in 2016. To obtain more information about the accuracy behaviour of the latest generation of hand-held 3D scanning systems, HCU Hamburg conducted further comparative geometrical investigations using structured light systems with speckle pattern (Artec Spider, Mantis Vision PocketScan 3D, Mantis Vision F5-SR, Mantis Vision F5-B, and Mantis Vision F6), and photogrammetric systems (Creaform HandySCAN 700 and Shining FreeScan X7). In the framework of these comparative investigations geometrically stable reference bodies were used. The appropriate reference data was acquired by measurements with two structured light projection systems (AICON smartSCAN and GOM ATOS I 2M). The comprehensive test results of the different test scenarios are presented and critically discussed in this contribution.

  8. A hand-held sensor for analyses of local distributions of magnetic fields and losses

    CERN Document Server

    Krismanic, G; Baumgartinger, N

    2000-01-01

    The paper describes a novel sensor for non-destructive analyses of local field and loss distributions in laminated soft magnetic cores, such as transformer cores. It was designed for rapid information on comparative local degrees of inhomogeneity, e.g., for the estimation of local building factors. Similar to a magnifying glass with handle, the compact hand-held sensor contains extremely sharp needle electrodes for the detection of the induction vector B as well as double-field coils for the vector H. Losses P are derived from the Poynting law. Applied to inner -- or also outer -- core regions, the sensor yields instantaneous computer displays of local H, B, and P.

  9. Balancing fast-rotating parts of hand-held machine drive

    Science.gov (United States)

    Korotkov, V. S.; Sicora, E. A.; Nadeina, L. V.; Yongzheng, Wang

    2018-03-01

    The article considers the issues related to the balancing of fast rotating parts of the hand-held machine drive including a wave transmission with intermediate rolling elements, which is constructed on the basis of the single-phase collector motor with a useful power of 1 kW and a nominal rotation frequency of 15000 rpm. The forms of balancers and their location are chosen. The method of balancing is described. The scheme for determining of residual unbalance in two correction planes is presented. Measurement results are given in tables.

  10. Center for X-Ray Optics, 1992

    International Nuclear Information System (INIS)

    1993-08-01

    This report discusses the following topics: Center for X-Ray Optics; Soft X-Ray Imaging wit Zone Plate Lenses; Biological X-Ray microscopy; Extreme Ultraviolet Lithography for Nanoelectronic Pattern Transfer; Multilayer Reflective Optics; EUV/Soft X-ray Reflectometer; Photoemission Microscopy with Reflective Optics; Spectroscopy with Soft X-Rays; Hard X-Ray Microprobe; Coronary Angiography; and Atomic Scattering Factors

  11. Exploring field-of-view non-uniformities produced by a hand-held spectroradiometer

    Directory of Open Access Journals (Sweden)

    Tamir Caras

    2011-01-01

    Full Text Available The shape of a spectroradiometer’s field of view (FOV affects the way spectral measurements are acquired. Knowing this property is a prerequisite for the correct use of the spectrometer. If the substrate is heterogeneous, the ability to accurately know what is being measured depends on knowing the FOV location, shape, spectral and spatial sensitivity. The GER1500 is a hand-held spectrometer with a fixed lens light entry slit and has a laser guide that allows control over the target by positioning the entire unit. In the current study, the FOV of the GER1500 was mapped and analysed. The spectral and spatial non-uniformities of the FOV were examined and were found to be spectrally independent. The relationship between the FOV and the built-in laser guide was tested and found to have a linear displacement dependent on the distance to the target. This allows an accurate prediction of the actual FOV position. A correction method to improve the agreement between the expected and measured reflectance over heterogeneous targets was developed and validated. The methods described are applicable and may be of use with other hand-held spectroradiometers.

  12. Neurosurgical hand-held optical coherence tomography (OCT) forward-viewing probe

    Science.gov (United States)

    Sun, Cuiru; Lee, Kenneth K. C.; Vuong, Barry; Cusimano, Michael; Brukson, Alexander; Mariampillai, Adrian; Standish, Beau A.; Yang, Victor X. D.

    2012-02-01

    A prototype neurosurgical hand-held optical coherence tomography (OCT) imaging probe has been developed to provide micron resolution cross-sectional images of subsurface tissue during open surgery. This new ergonomic hand-held probe has been designed based on our group's previous work on electrostatically driven optical fibers. It has been packaged into a catheter probe in the familiar form factor of the clinically accepted Bayonet shaped neurosurgical non-imaging Doppler ultrasound probes. The optical design was optimized using ZEMAX simulation. Optical properties of the probe were tested to yield an ~20 um spot size, 5 mm working distance and a 3.5 mm field of view. The scan frequency can be increased or decreased by changing the applied voltage. Typically a scan frequency of less than 60Hz is chosen to keep the applied voltage to less than 2000V. The axial resolution of the probe was ~15 um (in air) as determined by the OCT system. A custom-triggering methodology has been developed to provide continuous stable imaging, which is crucial for clinical utility. Feasibility of this probe, in combination with a 1310 nm swept source OCT system was tested and images are presented to highlight the usefulness of such a forward viewing handheld OCT imaging probe. Knowledge gained from this research will lay the foundation for developing new OCT technologies for endovascular management of cerebral aneurysms and transsphenoidal neuroendoscopic treatment of pituitary tumors.

  13. Standard guide to In-Plant performance evaluation of Hand-Held SNM monitors

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    1999-01-01

    1.1 This guide is one of a series on the application and evaluation of special nuclear material (SNM) monitors. Other guides in the series are listed in Section 2, and the relationship of in-plant performance evaluation to other procedures described in the series is illustrated in Fig. 1. Hand-held SNM monitors are described in of Guide C1112, and performance criteria illustrating their capabilities can be found in Appendix X1. 1.2 The purpose of this guide to in-plant performance evaluation is to provide a comparatively rapid procedure to verify that a hand-held SNM monitor performs as expected for detecting SNM or alternative test sources or to disclose the need for repair. The procedure can be used as a routine operational evaluation or it can be used to verify performance after a monitor is calibrated. 1.3 In-plant performance evaluations are more comprehensive than daily functional tests. They take place less often, at intervals ranging from weekly to once every three months, and derive their result fr...

  14. Hand-Held Sunphotometers for High School Student Construction and Measuring Aerosol Optical Thickness

    Science.gov (United States)

    Almonor, Linda; Baldwin, C.; Craig, R.; Johnson, L. P.

    2000-01-01

    Science education is taking the teaching of science from a traditional (lecture) approach to a multidimensional sense-making approach which allows teachers to support students by providing exploratory experiences. Using projects is one way of providing students with opportunities to observe and participate in sense-making activity. We created a learning environment that fostered inquiry-based learning. Students were engaged in a variety of Inquiry activities that enabled them to work in cooperative planning teams where respect for each other was encouraged and their ability to grasp, transform and transfer information was enhanced. Summer, 1998: An air pollution workshop was conducted for high school students in the Medgar Evers College/Middle College High School Liberty Partnership Summer Program. Students learned the basics of meteorology: structure and composition of the atmosphere and the processes that cause weather. The highlight of this workshop was the building of hand-held sunphotometers, which measure the intensity of the sunlight striking the Earth. Summer, 1999: high school students conducted a research project which measured the mass and size of ambient particulates and enhanced our ability to observe through land based measurements changes in the optical depth of ambient aerosols over Brooklyn. Students used hand held Sunphotometers to collect data over a two week period and entered it into the NASA GISS database by way of the internet.

  15. Code of safe practice for the use of x-rays in dentistry

    International Nuclear Information System (INIS)

    1991-06-01

    The purpose of this Code of Safe Practice for the Use of X-rays in Dentistry is to provide criteria for working procedures, x-ray equipment and protective materials necessary for the use of x-rays in dental diagnosis according to currently accepted standards of safety. Conformity with this code may be taken as a primary indication of compliance with radiation protection legislation. 7 refs., 1 ill

  16. Transmission X-ray mirror

    International Nuclear Information System (INIS)

    Lairson, B.M.; Bilderback, D.H.

    1982-01-01

    Transmission X-ray mirrors have been made from 400 A to 10 000 A thick soap films and have been shown to have novel properties. Using grazing angles of incidence, low energy X-rays were reflected from the front surface while more energetic X-rays were transmitted through the mirror largely unattenuated. A wide bandpass monochromator was made from a silicon carbide mirror followed by a soap film transmission mirror and operated in the white beam at the cornell High Energy Synchrotron Source (CHESS). Bandpasses of ΔE/E=12% to 18% were achieved at 13 keV with peak efficiencies estimated to be between 55% and 75%, respectively. Several wide angle scattering photographs of stretched polyethylene and a phospholipid were obtained in 10 s using an 18% bandpass. (orig.)

  17. Exponential x-ray transform

    International Nuclear Information System (INIS)

    Hazou, I.A.

    1986-01-01

    In emission computed tomography one wants to determine the location and intensity of radiation emitted by sources in the presence of an attenuating medium. If the attenuation is known everywhere and equals a constant α in a convex neighborhood of the support of f, then the problem reduces to that of inverting the exponential x-ray transform P/sub α/. The exponential x-ray transform P/sub μ/ with the attenuation μ variable, is of interest mathematically. For the exponential x-ray transform in two dimensions, it is shown that for a large class of approximate δ functions E, convolution kernels K exist for use in the convolution backprojection algorithm. For the case where the attenuation is constant, exact formulas are derived for calculating the convolution kernels from radial point spread functions. From these an exact inversion formula for the constantly attenuated transform is obtained

  18. X-ray of osteopathies

    International Nuclear Information System (INIS)

    Freyschmidt, J.

    1980-01-01

    Osteoporosis, osteomalcia, fibro-osteoclasia and osteosclerosis are essential reactions to pathologicometabolic processes of the bone. The X-ray film shows precisely which changes have taken place in the bone structure, thus supplying the means for an analysis based on anatomic pathology. These phenomena are discussed in detail, special attention being paid to structural modifications. Attention is also focused on the problems connected with X-ray technology. The value of direct and indirect magnification of the skeleton of the hand for the identification and classification of esteopathies is explained. Phenomena observed in X-ray films, such as enosteal erosion, intracortical longitudinal stripes or tunnelisation, as well as subperiostal absorption, can be of pathognomonic importance for certain osteopathies. (orig.) [de

  19. Imaging X-ray astronomy

    International Nuclear Information System (INIS)

    Elvis, M.

    1990-01-01

    The launch of the High Energy Astrophysical Observatory, more appealingly called the Einstein Observatory, marked one of the most revolutionary steps taken in astrophysics this century. Its greater sensitivity compared with earlier satellites and its ability to make high spacial and spectral resolution observations transformed X-ray astronomy. This book is based on a Symposium held in Cambridge, Massachusetts, to celebrate a decade of Einstein Observatory's achievements. It discusses the contributions that this satellite has made to each area of modern astrophysics and the diversity of the ongoing work based on Einstein data. There is a guide to each of the main data bases now coming on-line to increase the availability and to preserve this valuable archive for the future. A review of NASA's next big X-ray mission, AXAF, and a visionary program for novel X-ray astronomy satellites by Riccardo Giacconi conclude this wide-ranging volume. (author)

  20. Solar X-ray bursts

    International Nuclear Information System (INIS)

    Urnov, A.M.

    1980-01-01

    In the popular form the consideration is given to the modern state tasks and results of X-ray spectrometry of solar bursts. The operation of X-ray spectroheliograph is described. Results of spectral and polarization measurings of X-ray radiation of one powerful solar burst are presented. The conclusion has been drawn that in the process of burst development three characteristic stages may be distingwished: 1) the initial phase; just in this period processes which lead to observed consequences-electromagnetic and corpuscular radiation are born; 2) the impulse phase, or the phase of maximum, is characterised by sharp increase of radiation flux. During this phase the main energy content emanates and some volumes of plasma warm up to high temperatures; 3) the phase of burst damping, during which plasma cools and reverts to the initial condition

  1. X-ray fluorescence holography.

    Science.gov (United States)

    Hayashi, Kouichi; Happo, Naohisa; Hosokawa, Shinya; Hu, Wen; Matsushita, Tomohiro

    2012-03-07

    X-ray fluorescence holography (XFH) is a method of atomic resolution holography which utilizes fluorescing atoms as a wave source or a monitor of the interference field within a crystal sample. It provides three-dimensional atomic images around a specified element and has a range of up to a few nm in real space. Because of this feature, XFH is expected to be used for medium-range local structural analysis, which cannot be performed by x-ray diffraction or x-ray absorption fine structure analysis. In this article, we explain the theory of XFH including solutions to the twin-image problem, an advanced measuring system, and data processing for the reconstruction of atomic images. Then, we briefly introduce our recent applications of this technique to the analysis of local lattice distortions in mixed crystals and nanometer-size clusters appearing in the low-temperature phase of a shape-memory alloy.

  2. X-ray fluorescence holography

    International Nuclear Information System (INIS)

    Hayashi, Kouichi; Happo, Naohisa; Hosokawa, Shinya; Hu Wen; Matsushita, Tomohiro

    2012-01-01

    X-ray fluorescence holography (XFH) is a method of atomic resolution holography which utilizes fluorescing atoms as a wave source or a monitor of the interference field within a crystal sample. It provides three-dimensional atomic images around a specified element and has a range of up to a few nm in real space. Because of this feature, XFH is expected to be used for medium-range local structural analysis, which cannot be performed by x-ray diffraction or x-ray absorption fine structure analysis. In this article, we explain the theory of XFH including solutions to the twin-image problem, an advanced measuring system, and data processing for the reconstruction of atomic images. Then, we briefly introduce our recent applications of this technique to the analysis of local lattice distortions in mixed crystals and nanometer-size clusters appearing in the low-temperature phase of a shape-memory alloy. (topical review)

  3. Digital X-ray Imaging in Dentistry

    International Nuclear Information System (INIS)

    Kim, Eun Kyung

    1999-01-01

    In dentistry, Radio Visio Graphy was introduced as a first electronic dental x-ray imaging modality in 1989. Thereafter, many types of direct digital radiographic systems have been produced in the last decade. They are based either on charge-coupled device (CCD) or on storage phosphor technology. In addition, new types of digital radiographic system using amorphous selenium, image intensifier etc. are under development. Advantages of digital radiographic system are elimination of chemical processing, reduction in radiation dose, image processing, computer storage, electronic transfer of images and so on. Image processing includes image enhancement, image reconstruction, digital subtraction, etc. Especially digital subtraction and reconstruction can be applied in many aspects of clinical practice and research. Electronic transfer of images enables filmless dental hospital and teleradiology/teledentistry system. Since the first image management and communications system (IMACS) for dentomaxillofacial radiology was reported in 1992, IMACS in dental hospital has been increasing. Meanwhile, researches about computer-assisted diagnosis, such as structural analysis of bone trabecular patterns of mandible, feature extraction, automated identification of normal landmarks on cephalometric radiograph and automated image analysis for caries or periodontitis, have been performed actively in the last decade. Further developments in digital radiographic imaging modalities, image transmission system, imaging processing and automated analysis software will change the traditional clinical dental practice in the 21st century.

  4. Digital X-ray Imaging in Dentistry

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Eun Kyung [Dept. of Oral and Maxillofacial Radiology, College of Dentistry, Dankook University, Yongin (Korea, Republic of)

    1999-08-15

    In dentistry, Radio Visio Graphy was introduced as a first electronic dental x-ray imaging modality in 1989. Thereafter, many types of direct digital radiographic systems have been produced in the last decade. They are based either on charge-coupled device (CCD) or on storage phosphor technology. In addition, new types of digital radiographic system using amorphous selenium, image intensifier etc. are under development. Advantages of digital radiographic system are elimination of chemical processing, reduction in radiation dose, image processing, computer storage, electronic transfer of images and so on. Image processing includes image enhancement, image reconstruction, digital subtraction, etc. Especially digital subtraction and reconstruction can be applied in many aspects of clinical practice and research. Electronic transfer of images enables filmless dental hospital and teleradiology/teledentistry system. Since the first image management and communications system (IMACS) for dentomaxillofacial radiology was reported in 1992, IMACS in dental hospital has been increasing. Meanwhile, researches about computer-assisted diagnosis, such as structural analysis of bone trabecular patterns of mandible, feature extraction, automated identification of normal landmarks on cephalometric radiograph and automated image analysis for caries or periodontitis, have been performed actively in the last decade. Further developments in digital radiographic imaging modalities, image transmission system, imaging processing and automated analysis software will change the traditional clinical dental practice in the 21st century.

  5. Analysis of Information Remaining on Hand Held Devices Offered for Sale on the Second Hand Market

    Directory of Open Access Journals (Sweden)

    Andy Jones

    2008-06-01

    Full Text Available The ownership and use of mobile phones, Personal Digital Assistants and other hand held devices is now ubiquitous both for home and business use. The majority of these devices have a high initial cost, a relatively short period before they become obsolescent and a relatively low second hand value.  As a result of this, when the devices are replaced, there are indications that they tend to be discarded.  As technology has continued to develop, it has led to an increasing diversity in the number and type of devices that are available, and the processing power and the storage capacity of the digital storage in the device. All organisations, whether in the public or private sector increasingly use hand held devices that contain digital media for the storage of information relating to their business, their employees or their customers. Similarly, individual private users increasingly use hand held devices containing digital media for the storage of information relating to their private lives.The research revealed that a significant number of organisations and private users are ignorant or misinformed about the volume and type of information that is stored on the hand held devices and the media on which it is stored.  It is apparent that they have either not considered, or are unaware of, the potential impact of this information becoming available to their competitors or those with criminal intent.This main purpose of this study was to gain an understanding of the volume and type of information that may remain on hand held devices that are offered for sale on the second hand market.  A second aim of the research was to determine the level of damage that could, potentially be caused, if the information that remains on the devices fell into the wrong hands.  The study examined a number of hand held devices that had been obtained from sources in the UK and Australia that ranged from internet auction sites, to private sales and commercial

  6. Overutilization of x-rays

    International Nuclear Information System (INIS)

    Abrams, H.L.

    1979-01-01

    In this article on the overutilization of x-rays the author defines the term overutilization as excessive irradiation per unit of diagnostic information, therapeutic impact, or health outcome. Three main factors are described which lead to overutilization of x-rays: excessive radiation per film; excessive films per examination; and excessive examinations per patient. Topics discussed which influence the excessive examinations per patient are: the physician's lack of knowledge; undue dependence; lack of screening by radiologists; the physician's need for action and certainty; patient demand; reimbursement policies; institutional requirements; preventive medicine; defensive medicine; and the practice of radiology by nonradiologists

  7. Multichannel X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Khabakhpashev, A

    1980-10-01

    A typical design is discussed of multiwire proportional counters and their characteristic feature is explained, ie., the possibility of showing one or two coordinates of the X-ray quantum absorption site. The advantages of such instruments are listed, such as increased sensitivity of determination, the possibility of recording radiations of a different intensity, the possibility of on-line data processing and of the digital display of results. The fields of application include X-ray structural analysis in solid state physics, crystallography, molecular biology, astronomy, materials testing, and medicine.

  8. Semiconductor X-ray spectrometers

    International Nuclear Information System (INIS)

    Muggleton, A.H.F.

    1978-02-01

    An outline is given of recent developments in particle and photon induced x-ray fluorescence (XRF) analysis. Following a brief description of the basic mechanism of semiconductor detector operation a comparison is made between semiconductor detectors, scintillators and gas filled proportional devices. Detector fabrication and cryostat design are described in more detail and the effects of various device parameters on system performance, such as energy resolution, count rate capability, efficiency, microphony, etc. are discussed. The main applications of these detectors in x-ray fluorescence analysis, electron microprobe analysis, medical and pollution studies are reviewed

  9. Portable X-Ray Device

    Science.gov (United States)

    1983-01-01

    Portable x-ray instrument developed by NASA now being produced commercially as an industrial tool may soon find further utility as a medical system. The instrument is Lixiscope - Low Intensity X-Ray Imaging Scope -- a self-contained, battery-powered fluoroscope that produces an instant image through use of a small amount of radioactive isotope. Originally developed by Goddard Space Flight Center, Lixiscope is now being produced by Lixi, Inc. which has an exclusive NASA license for one version of the device.

  10. X-ray tube targets

    International Nuclear Information System (INIS)

    Hirsch, H.H.

    1980-01-01

    In rotary targets for X-ray tubes warping is a problem which causes X-ray deficiency. A rotary target is described in which warping is reduced by using alloys of molybdenum with 0.05 to 10% iron, silicon, cobalt, tantalum, niobium, hafnium, stable metal oxide or mixture thereof. Suitable mixtures are 0.5 to 10% of tantalum, niobium or hafnium with from 0.5 to 5% yttrium oxide, or 0.05 to 0.3% of cobalt or silicon. Optionally 0.1 to 5% by weight of additional material may be alloyed with the molybdenum, such as tantalum or hafnium carbides. (author)

  11. X-ray data processing.

    Science.gov (United States)

    Powell, Harold R

    2017-10-31

    The method of molecular structure determination by X-ray crystallography is a little over a century old. The history is described briefly, along with developments in X-ray sources and detectors. The fundamental processes involved in measuring diffraction patterns on area detectors, i.e. autoindexing, refining crystal and detector parameters, integrating the reflections themselves and putting the resultant measurements on to a common scale are discussed, with particular reference to the most commonly used software in the field. © 2017 The Author(s).

  12. X-ray Emission from Solar Flares

    Indian Academy of Sciences (India)

    2016-01-27

    Jan 27, 2016 ... Solar flares; X-ray detectors; X-ray line emission and continuum; break energy; microflares. Abstract. Solar X-ray Spectrometer (SOXS), the first space-borne solar astronomy experiment of India was designed to improve our current understanding of X-ray emission from the Sun in general and solar flares in ...

  13. Chest X-Ray (Chest Radiography)

    Science.gov (United States)

    ... Resources Professions Site Index A-Z X-ray (Radiography) - Chest Chest x-ray uses a very small dose ... Radiography? What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic ...

  14. X-Ray Exam: Cervical Spine

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Cervical Spine KidsHealth / For Parents / X-Ray ... MRI): Lumbar Spine Broken Bones Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  15. X-Ray Exam: Neck (For Parents)

    Science.gov (United States)

    ... Staying Safe Videos for Educators Search English Español X-Ray Exam: Neck KidsHealth / For Parents / X-Ray Exam: ... Neck Enlarged Adenoids Croup Sinusitis Getting an X-ray (Video) X-Ray (Video) View more Partner Message About Us ...

  16. Illumination system for X-ray lithography

    International Nuclear Information System (INIS)

    Buckley, W.D.

    1989-01-01

    An X-ray lithography system is described, comprising: a point source of X-Ray radiation; a wafer plane disposed in spaced relation to the point source of X-Ray radiation; a mask disposed between the point source of X-Ray radiation and the wafer plane whereby X-Ray radiation from the point source of X-ray radiation passes through the mask to the water plane; and X-Ray absorbent means mounted between the point source of X-Ray radiation and the wafer plane, the X-Ray absorbent means being of quadratically absorption from maximum absorption at the center to minimum absorption at the edge so as to have a radial absorption gradient profile to compensate for radial flux variation of the X-Ray radiation

  17. Gamma-ray detectors for intelligent, hand-held radiation monitors

    International Nuclear Information System (INIS)

    Fehlau, P.E.

    1983-01-01

    Small radiation detectors based on HgI 2 , bismuth germanate (BGO), plastic, or NaI(Tl) detector materials were evaluated for use in small, lighweight radiation monitors. The two denser materials, HgI 2 and BGO, had poor resolution at low-energy and thus performed less well than NaI(Tl) in detecting low-energy gamma rays from bare, enriched uranium. The plastic scintillator, a Compton recoil detector, also performed less well at low gamma-ray energy. Two small NaI(Tl) detectors were suitable for detecting bare uranium and sheilded plutonium. One became part of a new lightweight hand-held monitor and the other found uses as a pole-mounted detector for monitoring hard-to-reach locations

  18. Hand-held dynamic visual noise reduces naturally occurring food cravings and craving-related consumption.

    Science.gov (United States)

    Kemps, Eva; Tiggemann, Marika

    2013-09-01

    This study demonstrated the applicability of the well-established laboratory task, dynamic visual noise, as a technique for reducing naturally occurring food cravings and subsequent food intake. Dynamic visual noise was delivered on a hand-held computer device. Its effects were assessed within the context of a diary study. Over a 4-week period, 48 undergraduate women recorded their food cravings and consumption. Following a 2-week baseline, half the participants watched the dynamic visual noise display whenever they experienced a food craving. Compared to a control group, these participants reported less intense cravings. They were also less likely to eat following a craving and consequently consumed fewer total calories following craving. These findings hold promise for curbing unwanted food cravings and craving-driven consumption in real-world settings. Copyright © 2013 Elsevier Ltd. All rights reserved.

  19. Design of hand held RID's monitoring system based on embedded system

    International Nuclear Information System (INIS)

    Wang Hongwei; Wei Yixiang

    2008-01-01

    In this paper we introduce the design of monitoring system for the hand held radionuclide identification device (RID), constructed under the embedded operating system of WinCE. At first, we introduce the design of hardware and software platform, and following is the major part of technical view of the software system, including the driver development, P/Invoke mechanism to call the C/C++ subroutines, multi-thread technology. In the experimental hardware platform, we have developed a front-end monitoring system for portable device targeted nuclide identification and orientation. It's a full-featured and flexible system, with the functions of data acquisition, radioactivity locating, data import and export, etc. (authors)

  20. A hand-held 3D laser scanning with global positioning system of subvoxel precision

    International Nuclear Information System (INIS)

    Arias, Nestor; Meneses, Nestor; Meneses, Jaime; Gharbi, Tijani

    2011-01-01

    In this paper we propose a hand-held 3D laser scanner composed of an optical head device to extract 3D local surface information and a stereo vision system with subvoxel precision to measure the position and orientation of the 3D optical head. The optical head is manually scanned over the surface object by the operator. The orientation and position of the 3D optical head is determined by a phase-sensitive method using a 2D regular intensity pattern. This phase reference pattern is rigidly fixed to the optical head and allows their 3D location with subvoxel precision in the observation field of the stereo vision system. The 3D resolution achieved by the stereo vision system is about 33 microns at 1.8 m with an observation field of 60cm x 60cm.

  1. The Complimentary Role of Methoxy-Isobutyl-Isonitrile and Hand-Held Gamma Probe in Adamantinoma

    Science.gov (United States)

    Maharaj, Masha; Korowlay, Nisaar; Ellmann, Prof

    2016-01-01

    Adamantinoma is a rare locally aggressive osteolytic tumor that is found 90% of the time in the diaphysis of the tibia with the remaining lesions found in the fibula and long tubular bones. A case of adamantinoma of the tibia is presented. The added value of nuclear medicine investigations in the workup of this patient is described. A three-phase whole body 99mTc-methylene diphosphonate bone and a whole body 99mTc-methoxy-isobutyl-isonitrile scans were complimentary in the demarcation of viable bone tumor and the assessment of the remainder of the bone and soft tissue to exclude other sites. Intra-operative assistance with a hand-held gamma probe, guided the biopsy of the most metabolically active tumor tissue. Histology revealed a biphasic tumor composed of epithelial and fibrous components, in keeping with an adamantinoma. PMID:26912979

  2. Clinical assessment of hip strength using a hand-held dynamometer is reliable

    DEFF Research Database (Denmark)

    Thorborg, K; Petersen, J; Magnusson, S P

    2010-01-01

    rotation (ER), internal rotation (IR), flexion (FLEX) and extension (EXT) using a hand-held dynamometer. Nine subjects (five males, four females), physically active for at least 2.5 h a week, were included. Twelve standardized isometric strength tests were performed twice with a 1-week interval in between......Hip strength assessment plays an important role in the clinical examination of the hip and groin region. The primary aim of this study was to examine the absolute test-retest measurement variation concerning standardized strength assessments of hip abduction (ABD), adduction (ADD), external...... by the same examiner. The test order was randomized to avoid systematic bias. Measurement variation between sessions was 3-12%. When the maximum value of four measurements was used, test-retest measurement variation was below 10% in 11 of the 12 individual hip strength tests and below 5% in five of the 12...

  3. Implementation of synthetic aperture imaging on a hand-held device

    DEFF Research Database (Denmark)

    Hemmsen, Martin Christian; Kjeldsen, Thomas; Larsen, Lee

    2014-01-01

    -held devices all with different chipsets and a BK Medical UltraView 800 ultrasound scanner emulating a wireless probe. The wireless transmission is benchmarked using an imaging setup consisting of 269 scan lines x 1472 complex samples (1.58 MB pr. frame, 16 frames per second). The measured data throughput...... reached an average of 28.8 MB/s using a LG G2 mobile device, which is more than the required data throughput of 25.3 MB/s. Benchmarking the processing performance for B-mode imaging showed a total processing time of 18.9 ms (53 frames/s), which is less than the acquisition time (62.5 ms).......This paper presents several implementations of Syn- thetic Aperture Sequential Beamforming (SASB) on commer- cially available hand-held devices. The implementations include real-time wireless reception of ultrasound radio frequency sig- nals and GPU processing for B-mode imaging. The proposed...

  4. Digital knowledge in the coat pocket - hand-held personal digital assistants in radiology

    International Nuclear Information System (INIS)

    Niehues, S.M.; Froehlich, M.; Felix, R.; Lemke, A.J.

    2004-01-01

    The personal digital assistant (PDA) enables the independent access to large data in a pocket-sized format. The applications for hand-held computers are growing steadily and can support almost any kind of problem. An overview of the available hardware and software is provided and evaluated. Furthermore, the use of the PDA in the clinical daily routine is described. In view of the numerous software programs available in radiology, the range of software solutions for radiologists is presented. Despite the high acquisition cost, the PDA has already become the digital assistant for the radiologist. After a short time of getting used to the PDA, nobody wants to miss it at work or at home. New technical features and available software programs will continuously increase the integration of the PDA into the medical workflow in the near future. (orig.)

  5. Cost effective spectral sensor solutions for hand held and field applications

    Science.gov (United States)

    Reetz, Edgar; Correns, Martin; Notni, Gunther

    2015-05-01

    Optical spectroscopy is without doubt one of the most important non-contact measurement principles. It is used in a wide range of applications from bio-medical to industrial fields. One recent trend is to miniaturize spectral sensors to address new areas of application. The most common spectral sensor type is based on diffraction gratings, while other types are based on micro mechanical systems (MEMS) or filter technologies. The authors represent the opinion that there is a potentially wide spread field of applications for spectrometers, but the market limits the range of applications since they cannot keep up with targeted cost requirements for consumer products. The present article explains an alternative approach for miniature multichannel spectrometer to enhance robustness for hand held field applications at a cost efficient price point.

  6. X-ray hot plasma diagnostics

    International Nuclear Information System (INIS)

    Cojocaru, E.

    1984-11-01

    X-ray plasma emission study is powerful diagnostic tool of hot plasmas. In this review article the main techniques of X-ray plasma emission measurement are shortly presented: X-ray spectrometry using absorbent filters, crystal and grating spectrometers, imaging techniques using pinhole cameras, X-ray microscopes and Fresnel zone plate cameras, X-ray plasma emission calorimetry. Advances in these techniques with examples for different hot plasma devices are also presentes. (author)

  7. An integrated portable hand-held analyser for real-time isothermal nucleic acid amplification

    Energy Technology Data Exchange (ETDEWEB)

    Smith, Matthew C. [College of Marine Science, University of South Florida, St Petersburg, FL (United States)], E-mail: msmith@marine.usf.edu; Steimle, George; Ivanov, Stan; Holly, Mark; Fries, David P. [College of Marine Science, University of South Florida, St Petersburg, FL (United States)

    2007-08-29

    A compact hand-held heated fluorometric instrument for performing real-time isothermal nucleic acid amplification and detection is described. The optoelectronic instrument combines a Printed Circuit Board/Micro Electro Mechanical Systems (PCB/MEMS) reaction detection/chamber containing an integrated resistive heater with attached miniature LED light source and photo-detector and a disposable glass waveguide capillary to enable a mini-fluorometer. The fluorometer is fabricated and assembled in planar geometry, rolled into a tubular format and packaged with custom control electronics to form the hand-held reactor. Positive or negative results for each reaction are displayed to the user using an LED interface. Reaction data is stored in FLASH memory for retrieval via an in-built USB connection. Operating on one disposable 3 V lithium battery >12, 60 min reactions can be performed. Maximum dimensions of the system are 150 mm (h) x 48 mm (d) x 40 mm (w), the total instrument weight (with battery) is 140 g. The system produces comparable results to laboratory instrumentation when performing a real-time nucleic acid sequence-based amplification (NASBA) reaction, and also displayed comparable precision, accuracy and resolution to laboratory-based real-time nucleic acid amplification instrumentation. A good linear response (R{sup 2} = 0.948) to fluorescein gradients ranging from 0.5 to 10 {mu}M was also obtained from the instrument indicating that it may be utilized for other fluorometric assays. This instrument enables an inexpensive, compact approach to in-field genetic screening, providing results comparable to laboratory equipment with rapid user feedback as to the status of the reaction.

  8. Reliability of measuring hip abductor strength following total knee arthroplasty using a hand-held dynamometer.

    Science.gov (United States)

    Schache, Margaret B; McClelland, Jodie A; Webster, Kate E

    2016-01-01

    To investigate the test-retest reliability of measuring hip abductor strength in patients with total knee arthroplasty (TKA) using a hand-held dynamometer (HHD) with two different types of resistance: belt and manual resistance. Test-retest reliability of 30 subjects (17 female, 13 male, 71.9 ± 7.4 years old), 9.2 ± 2.7 days post TKA was measured using belt and therapist resistance. Retest reliability was calculated with intra-class coefficients (ICC3,1) and 95% confidence intervals (CI) for both the group average and the individual scores. A paired t-test assessed whether a difference existed between the belt and therapist methods of resistance. ICCs were 0.82 and 0.80 for the belt and therapist resisted methods, respectively. Hip abductor strength increases of 8 N (14%) for belt resisted and 14 N (17%) for therapist resisted measurements of the group average exceeded the 95% CI and may represent real change. For individuals, hip abductor strength increases of 33 N (72%) (belt resisted) and 57 N (79%) (therapist resisted) could be interpreted as real change. Hip abductor strength can be reliably measured using HHD in the clinical setting with the described protocol. Belt resistance demonstrated slightly higher test-retest reliability. Reliable measurement of hip abductor muscle strength in patients with TKA is important to ensure deficiencies are addressed in rehabilitation programs and function is maximized. Hip abductor strength can be reliably measured with a hand-held dynamometer in the clinical setting using manual or belt resistance.

  9. An integrated portable hand-held analyser for real-time isothermal nucleic acid amplification

    International Nuclear Information System (INIS)

    Smith, Matthew C.; Steimle, George; Ivanov, Stan; Holly, Mark; Fries, David P.

    2007-01-01

    A compact hand-held heated fluorometric instrument for performing real-time isothermal nucleic acid amplification and detection is described. The optoelectronic instrument combines a Printed Circuit Board/Micro Electro Mechanical Systems (PCB/MEMS) reaction detection/chamber containing an integrated resistive heater with attached miniature LED light source and photo-detector and a disposable glass waveguide capillary to enable a mini-fluorometer. The fluorometer is fabricated and assembled in planar geometry, rolled into a tubular format and packaged with custom control electronics to form the hand-held reactor. Positive or negative results for each reaction are displayed to the user using an LED interface. Reaction data is stored in FLASH memory for retrieval via an in-built USB connection. Operating on one disposable 3 V lithium battery >12, 60 min reactions can be performed. Maximum dimensions of the system are 150 mm (h) x 48 mm (d) x 40 mm (w), the total instrument weight (with battery) is 140 g. The system produces comparable results to laboratory instrumentation when performing a real-time nucleic acid sequence-based amplification (NASBA) reaction, and also displayed comparable precision, accuracy and resolution to laboratory-based real-time nucleic acid amplification instrumentation. A good linear response (R 2 = 0.948) to fluorescein gradients ranging from 0.5 to 10 μM was also obtained from the instrument indicating that it may be utilized for other fluorometric assays. This instrument enables an inexpensive, compact approach to in-field genetic screening, providing results comparable to laboratory equipment with rapid user feedback as to the status of the reaction

  10. Combining heterogenous features for 3D hand-held object recognition

    Science.gov (United States)

    Lv, Xiong; Wang, Shuang; Li, Xiangyang; Jiang, Shuqiang

    2014-10-01

    Object recognition has wide applications in the area of human-machine interaction and multimedia retrieval. However, due to the problem of visual polysemous and concept polymorphism, it is still a great challenge to obtain reliable recognition result for the 2D images. Recently, with the emergence and easy availability of RGB-D equipment such as Kinect, this challenge could be relieved because the depth channel could bring more information. A very special and important case of object recognition is hand-held object recognition, as hand is a straight and natural way for both human-human interaction and human-machine interaction. In this paper, we study the problem of 3D object recognition by combining heterogenous features with different modalities and extraction techniques. For hand-craft feature, although it reserves the low-level information such as shape and color, it has shown weakness in representing hiconvolutionalgh-level semantic information compared with the automatic learned feature, especially deep feature. Deep feature has shown its great advantages in large scale dataset recognition but is not always robust to rotation or scale variance compared with hand-craft feature. In this paper, we propose a method to combine hand-craft point cloud features and deep learned features in RGB and depth channle. First, hand-held object segmentation is implemented by using depth cues and human skeleton information. Second, we combine the extracted hetegerogenous 3D features in different stages using linear concatenation and multiple kernel learning (MKL). Then a training model is used to recognize 3D handheld objects. Experimental results validate the effectiveness and gerneralization ability of the proposed method.

  11. X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Vries, J.L. de.

    1976-01-01

    The seventh edition of Philips' Review of Literature on x-ray fluorescence spectrometry starts with a list of conference proceedings on the subject, organised by the Philips organisation at regular intervals in various European countries. It is followed by a list of bulletins. The bibliography is subdivided according to spectra, equipment, applications and absorption analysis

  12. Stellar X-Ray Polarimetry

    Science.gov (United States)

    Swank, J.

    2011-01-01

    Most of the stellar end-state black holes, pulsars, and white dwarfs that are X-ray sources should have polarized X-ray fluxes. The degree will depend on the relative contributions of the unresolved structures. Fluxes from accretion disks and accretion disk corona may be polarized by scattering. Beams and jets may have contributions of polarized emission in strong magnetic fields. The Gravity and Extreme Magnetism Small Explorer (GEMS) will study the effects on polarization of strong gravity of black holes and strong magnetism of neutron stars. Some part of the flux from compact stars accreting from companion stars has been reflected from the companion, its wind, or accretion streams. Polarization of this component is a potential tool for studying the structure of the gas in these binary systems. Polarization due to scattering can also be present in X-ray emission from white dwarf binaries and binary normal stars such as RS CVn stars and colliding wind sources like Eta Car. Normal late type stars may have polarized flux from coronal flares. But X-ray polarization sensitivity is not at the level needed for single early type stars.

  13. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available Toggle navigation Test/Treatment Patient Type Screening/Wellness Disease/Condition Safety En Español More Info Images/Videos About Us News Physician Resources Professions Site Index A-Z X-ray (Radiography) - Bone ...

  14. X-ray Sensitive Material

    Science.gov (United States)

    2015-12-01

    TM 8772 5 Literature reports on BiI3/nylon composites used X-ray sources with a Mo target (Reference 18) or magnesium target (Reference 19). However...1981. Pp. D-155 to D-160. 22. G. Pretzsch, B. Dorschel, and T. Schonmuth. IEEE Transactions on Electrical Insulation, Vol. EI -21, No.3, June 1986

  15. X-ray system analysis

    International Nuclear Information System (INIS)

    Shapiro, J.S.

    1985-01-01

    An X-ray system tester for measuring anode voltage, cathode voltage, anode current, filament current and line voltage in an X-ray system has a selector which couples one of these analog signals or one of a plurality of processing control signals entered by an operator from a control panel to a digitizing section selectively in accordance with control signals provided to the selector by a computing section. The digitizing section converts the selected signal into a train of pulses having a frequency proportional to the value of the selected signal. These pulses are counted, the counts being used by the computing section to determine the value of the selected signal. This computed value is stored in a computing memory section of the computing section. The computing section is adapted to store a plurality of the sets of signals produced during a corresponding sequence of operational intervals of the X-ray system and determines a measure of the deviation of any selected one of the stored electrical signals over the sequence of operating intervals. Each signal produced during the sequential operational intervals can be recalled to aid analysis of the operation of the X-ray system. (author)

  16. X-ray image coding

    International Nuclear Information System (INIS)

    1974-01-01

    The invention aims at decreasing the effect of stray radiation in X-ray images. This is achieved by putting a plate between source and object with parallel zones of alternating high and low absorption coefficients for X-radiation. The image is scanned with the help of electronic circuits which decode the signal space coded by the plate, thus removing the stray radiation

  17. X-ray simulation development

    International Nuclear Information System (INIS)

    Posey, L.D.; Tollefsrud, P.B.; Woodall, H.W.; McDaniel, D.H.; Allred, R.E.

    1975-01-01

    Design modifications are discussed for an electron beam accelerator used as a Bremsstrahlung x-ray source. The primary goal of the program, to obtain a reliable 5 cal/gm exposure capability, can be accomplished with beam compression by an external magnetic guide field. Initial operating characteristics and performance improvements are presented

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... ray examination. X-rays usually have no side effects in the typical diagnostic range for this exam. Risks There is always a slight chance of cancer from excessive exposure to radiation. However, the benefit of an accurate ...

  19. X-ray fluorescence spectrometry

    International Nuclear Information System (INIS)

    Ray, N.B.

    1977-01-01

    The principle, instrument and procedure of X-ray fluorescence spectrometry are described. It is a rapid, simple and sensitive method for the trace analysis of elements from sodium to uranium in powder, liquid or metal samples. (M.G.B.)

  20. X-rays from stars

    Science.gov (United States)

    Güdel, Manuel

    2004-07-01

    Spectroscopic studies available from Chandra and XMM-Newton play a pivotal part in the understanding of the physical processes in stellar (magnetic and non-magnetic) atmospheres. It is now routinely possible to derive densities and to study the influence of ultraviolet radiation fields, both of which can be used to infer the geometry of the radiating sources. Line profiles provide important information on bulk mass motions and attenuation by neutral matter, e.g. in stellar winds. The increased sensitivity has revealed new types of X-ray sources in systems that were thought to be unlikely places for X-rays: flaring brown dwarfs, including rather old, non-accreting objects, and terminal shocks in jets of young stars are important examples. New clues concerning the role of stellar high-energy processes in the modification of the stellar environment (ionization, spallation, etc.) contribute significantly to our understanding of the "astro-ecology" in forming planetary systems. Technological limitations are evident. The spectral resolution has not reached the level where bulk mass motions in cool stars become easily measurable. Higher resolution would also be important to perform X-ray "Doppler imaging" in order to reconstruct the 3-D distribution of the X-ray sources around a rotating star. Higher sensitivity will be required to perform high-resolution spectroscopy of weak sources such as brown dwarfs or embedded pre-main-sequence sources. A new generation of satellites such as Constellation-X or XEUS should pursue these goals.

  1. Chandra's X-ray Vision

    Indian Academy of Sciences (India)

    1999-07-23

    Jul 23, 1999 ... CXO is 13.8 metres long and its solar arrays have a wingspan of. 19.5 metres as shown in ... the Universe (for example, coronae of stars, matter ejected from .... The telescope system and the scientific instruments were put through ..... solve the puzzle about the origin of cosmic X-ray background- one of the ...

  2. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available Toggle navigation Test/Treatment Patient Type Screening/Wellness Disease/Condition Safety En Español More Info Images/Videos About Us News Physician ... An x-ray (radiograph) is a noninvasive medical test that helps physicians diagnose and treat medical conditions. ...

  3. Stellar X-ray sources

    International Nuclear Information System (INIS)

    Katz, J.I.; Washington Univ., St. Louis, MO

    1988-01-01

    I Review some of the salient accomplishments of X-rap studies of compact objects. Progress in this field has closely followed the improvement of observational methods, particularly in angular resolution and duration of exposure. Luminous compact X-ray sources are accreting neutron stars or black holes. Accreting neutron stars may have characteristic temporal signatures, but the only way to establish that an X-ray source is a black hole is to measure its mass. A rough phenomenological theory is succesful, but the transport of angular momentum in accretion flows is not onderstood. A number of interesting complications have been observed, including precessing accretion discs, X-ray bursts, and the acceleration of jets in SS433. Many puzzles remain unsolved, including the excitation of disc precession, the nature of the enigmatic A- and gamma-ray source Cyg X-3, the mechanism by which slowly spinning accreting neutron stars lose angular momentum, and the superabundance of X-ray sources in globular clusters. 41 refs.; 5 figs

  4. X-rays and magnetism

    International Nuclear Information System (INIS)

    Fischer, Peter; Ohldag, Hendrik

    2015-01-01

    Magnetism is among the most active and attractive areas in modern solid state physics because of intriguing phenomena interesting to fundamental research and a manifold of technological applications. State-of-the-art synthesis of advanced magnetic materials, e.g. in hybrid structures paves the way to new functionalities. To characterize modern magnetic materials and the associated magnetic phenomena, polarized x-rays have emerged as unique probes due to their specific interaction with magnetic materials. A large variety of spectroscopic and microscopic techniques have been developed to quantify in an element, valence and site-sensitive way properties of ferro-, ferri-, and antiferromagnetic systems, such as spin and orbital moments, and to image nanoscale spin textures and their dynamics with sub-ns time and almost 10 nm spatial resolution. The enormous intensity of x-rays and their degree of coherence at next generation x-ray facilities will open the fsec time window to magnetic studies addressing fundamental time scales in magnetism with nanometer spatial resolution. This review will give an introduction into contemporary topics of nanoscale magnetic materials and provide an overview of analytical spectroscopy and microscopy tools based on x-ray dichroism effects. Selected examples of current research will demonstrate the potential and future directions of these techniques. (report on progress)

  5. X-ray examination apparatus

    International Nuclear Information System (INIS)

    Bernstein, S.; Griswa, P.J.; Halter, P. Jr.; Kidd, H.J.

    1980-01-01

    Apparatus for x-ray cardiovascular examination and which can also be used for general purpose examination is described. An advantage of the system is that there is no mechanical connection between the image intensifier and source to interfere with the medical examiner or emergency procedures. (U.K.)

  6. X-ray tube transformer

    International Nuclear Information System (INIS)

    1980-01-01

    An X-ray generator is described which comprises a transmission line transformer including an electrical conductor with a cavity and a second electrical conductor including helical windings disposed along a longitudinal axis within the cavity of the first conductor. The windings have a pitch which varies per unit length along the axis. There is dielectric material in the cavity for insulation and to couple electromagnetically the two conductors in response to an electric current flowing through the conductors, which have an impedance between them; this varies with distance along the axis of the helix of the second conductor. An X-ray tube is disposed along the longitudinal axis within the cavity, for radiating X-rays. The invention increases the voltage of applied voltage pulses at the remote tube-head with a transformer formed by using a spiral delay line geometry to give a tapered-impedance coaxial high voltage multiplier for pulse voltage operation. This transformer is smaller and lighter than previous designs for the same high peak voltage and power ratings. This is important because the penetration capabilities of Flash X-ray equipment increase with voltage, particularly in heavy materials such as steel. (U.K.)

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... of bone cancer . locate foreign objects in soft tissues around or in bones. top of page How should I prepare? Most ... absorb the x-rays in varying degrees. Dense bone absorbs much of the radiation while soft tissue, such as muscle, fat and organs, allow more ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... navigation Test/Treatment Patient Type Screening/Wellness Disease/Condition Safety En Español More Info Images/Videos About Us ... the equipment look like? How does the procedure work? How is the procedure ... diagnose and treat medical conditions. Imaging with x-rays involves exposing a part ...

  9. Proton exciting X ray analysis

    International Nuclear Information System (INIS)

    Ma Xinpei

    1986-04-01

    The analyzing capability of proton exciting X ray analysis for different elements in organisms was discussed, and dealing with examples of trace element analysis in the human body and animal organisms, such as blood serum, urine, and hair. The sensitivity, accuracy, and capability of multielement analysis were discussed. Its strong points for the trace element analysis in biomedicine were explained

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... over time. top of page What are the benefits vs. risks? Benefits Bone x-rays are the fastest and easiest ... cancer from excessive exposure to radiation. However, the benefit of an accurate diagnosis far outweighs the risk. ...

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available Toggle navigation Test/Treatment Patient Type Screening/Wellness Disease/Condition Safety En Español More Info Images/Videos About Us News Physician Resources Professions Site Index A-Z X-ray ( ...

  12. X-ray absorption holography

    Czech Academy of Sciences Publication Activity Database

    Kopecký, Miloš; Lausi, A.; Bussetto, E.; Kub, Jiří; Savoia, A.

    2002-01-01

    Roč. 88, č. 18 (2002), s. 185503-1 - 185503-3 ISSN 0031-9007 R&D Projects: GA MŠk LN00A100 Institutional research plan: CEZ:AV0Z1010921 Keywords : x-ray holography Subject RIV: BH - Optics, Masers, Lasers Impact factor: 7.323, year: 2002

  13. Evaluation device of diagnostic X-ray beam

    International Nuclear Information System (INIS)

    Pela, Carlos Alberto; Paula, Eldereis de; Bruco, Jose Luiz; Ghilardi Netto, Thomaz

    1996-01-01

    A device developed to determine the physical parameters useful to the evaluation of both dental and medical X-ray equipment is presented. Nine detectors connected each one to a signal amplifier are used. The amplifiers output has being connected to an analog to digital interface plugged in a PC-AT microcomputer. The X-ray pass through filters before reaching the detectors. The data collected by microcomputer are used to calculate the kVp and other factors as total tube filtration, waveform, half-value layer, exposure rate and exposure time

  14. Measuring device for soft X-rays

    International Nuclear Information System (INIS)

    Dissing, E.

    1978-09-01

    An instrument for the measurement of the absorbed energy per unit area of diagnostic X-rays in soft human tissue was developed. The instrument is intended for dosimetry applications in the field of dental and small skeleton radiography and for mammography. The detector assembly consists of a Polyvinyltoluene scintillator 2.54 diametre x 5.08 cm CsSb semitransparent head-on vacuum phototube. Polyvinyltoluene being a pure hydrocarbon may be considered a good representative material of human soft tissue concerning the absorption of X-rays. In the photon energy range of interest, 5 - 40 keV, the mass energy absorption coefficient for muscle tissue and for PVT differ about a factor 2 due to the considerable content of Oxygen in muscle tissue. This is to some extend reflected in the photon energy response characteristic for the instrument. For human adipose, the characteristic is practically flat from 5- 40 keV. The instrument is integrating the absorbed power per unit area and the digital display shows Joules/m 2 . The range for the instrument is from 000.1 μJ/m 2 to 19.99 J/m 2 (absorbed energy in 5 cm tissue). (author)

  15. High-Resolution X-ray Emission and X-ray Absorption Spectroscopy

    NARCIS (Netherlands)

    Groot, F.M.F. de

    2000-01-01

    In this review, high-resolution X-ray emission and X-ray absorption spectroscopy will be discussed. The focus is on the 3d transition-metal systems. To understand high-resolution X-ray emission and reso-nant X-ray emission, it is first necessary to spend some time discussing the X-ray absorption

  16. Characterization of contaminated soils using a x-ray fluorescence equipment

    International Nuclear Information System (INIS)

    Quindos, L. S.; Fuente, I.; Sainz, C.; Queral, I.

    2006-01-01

    Fast and accurate material identification is required in many areas of the metal industry as well as surveillance of contaminated areas. The portable X-ray analyzer described in this paper consists of a hand-held probe and electronic unit. The system is configured with three excitation sources, such as ''241 Am, ''109Cd and ''55 Fe, which provides excellent performance for common elements to be controlled in contaminated soils. Data about measurements developed with this device in field as well as in the laboratory are shown in this paper. (Author)

  17. Early reports of x ray dangers

    International Nuclear Information System (INIS)

    Mazzola, P.V.

    1974-01-01

    The development of diagnostic radiography is one of the most valuable tools in the diagnosis of medical and dental problems today. This has only been achieved after its destructive and mutating effects on human tissue were recognized, and protective devices were adopted and exposure routines were developed that resulted in making a radiographic examination the relatively safe and simple procedure it is today. It is unfortunate that many of the earlier workers in this field suffered exposure to a degree that led to required amputation of hands and fingers, to malignancies at the sites of exposure, and, in several cases, even to death. But without these intrepid and earnest pioneers in an unchartered course, the tremendous boon to mankind which is the x ray may never have reached its full potential. (U.S.)

  18. Automatic segmentation of mandible in panoramic x-ray.

    Science.gov (United States)

    Abdi, Amir Hossein; Kasaei, Shohreh; Mehdizadeh, Mojdeh

    2015-10-01

    As the panoramic x-ray is the most common extraoral radiography in dentistry, segmentation of its anatomical structures facilitates diagnosis and registration of dental records. This study presents a fast and accurate method for automatic segmentation of mandible in panoramic x-rays. In the proposed four-step algorithm, a superior border is extracted through horizontal integral projections. A modified Canny edge detector accompanied by morphological operators extracts the inferior border of the mandible body. The exterior borders of ramuses are extracted through a contour tracing method based on the average model of mandible. The best-matched template is fetched from the atlas of mandibles to complete the contour of left and right processes. The algorithm was tested on a set of 95 panoramic x-rays. Evaluating the results against manual segmentations of three expert dentists showed that the method is robust. It achieved an average performance of [Formula: see text] in Dice similarity, specificity, and sensitivity.

  19. Center for X-Ray Optics, 1986

    International Nuclear Information System (INIS)

    1987-07-01

    The Center for X-Ray Optics has made substantial progress during the past year on the development of very high resolution x-ray technologies, the generation of coherent radiation at x-ray wavelengths, and, based on these new developments, had embarked on several scientific investigations that would not otherwise have been possible. The investigations covered in this report are topics on x-ray sources, x-ray imaging and applications, soft x-ray spectroscopy, synchrotron radiation, advanced light source and magnet structures for undulators and wigglers

  20. X-ray diagnostic in gas discharge

    International Nuclear Information System (INIS)

    Chen Suhe; Wang Dalun; Cui Gaoxian; Wang Mei; Fu Yibei; Zhang Xinwei; Zhang Wushou

    1995-01-01

    X rays were observed when the anomalous phenomenon in the metal loaded with deuterium studied by the gas-discharge method. Therefore the X-ray energy spectra were measured by the absorption method, the specific X-ray approach and the NaI scintillation counter, while X-ray intensity was estimated by using 7 Li thermoluminescent foils. The X-ray average energy measured by the absorption method is 27.6 +- 2.1 keV, which is fitted within the error extent to 26.0 +-2.4 keV monoenergetic X rays measured by the NaI scintillation counter

  1. X-ray cardiovascular examination apparatus

    International Nuclear Information System (INIS)

    1977-01-01

    An X-ray source is mounted in an enclosure for angulating longitudinally about a horizontal axis. An X-ray-permeable, patient-supporting table is mounted on the top of the enclosure for executing lateral and longitudinal movements. An X-ray image-receiving device such as an X-ray image intensifier is mounted above the table on a vertically movable arm which is on a longitudinally movable carriage. Electric control means are provided for angulating the X-ray source and image intensifier synchronously as the image intensifier system is shifted longitudinally or vertically such that the central ray from the X-ray source is kept intensifier

  2. X-ray tube arrangements

    International Nuclear Information System (INIS)

    Gillard, R.G.

    1980-01-01

    A technique for ensuring the rapid correction of both amplitude and offset errors in the deflectional movement of an electron beam along an X-ray emissive target is described. The movement is monitored at at least two positions during a sweep and differences, between the two movements and a desired movement, at these positions are combined in different proportions to produce a corrective servo signal. Such arrangements find application, for example, in computerised tomographic scanners. (author)

  3. Smart X-ray optics

    International Nuclear Information System (INIS)

    Michette, A G; Pfauntsch, S J; Sahraei, S; Shand, M; Morrison, G R; Hart, D; Vojnovic, B; Stevenson, T; Parkes, W; Dunare, C; Willingale, R; Feldman, C; Button, T; Zhang, D; Rodriguez-Sanmartin, D; Wang, H

    2009-01-01

    This paper describes reflective adaptive/active optics for applications including studies of biological radiation damage. The optics work on the polycapillary principle, but use arrays of channels in thin silicon. For optimum performance the x-rays should reflect once off a channel wall in each of two successive arrays. This reduces aberrations since then the Abbe sine condition is approximately satisfied. Adaptivity is achieved by flexing the arrays via piezo actuation, providing further aberration reduction and controllable focal length.

  4. Advanced imaging technology using carbon nanotube x ray source

    International Nuclear Information System (INIS)

    Choi, Hae Young; Seol, Seung Kown; Kim, Jaehoon; Yoo, Seung Hoon; Kim, Jong Uk

    2008-01-01

    Recently, X ray imaging technology is a useful and leading medical diagnostic tool for healthcare professionals to diagnose disease in human body. CNTs(i.e. carbon nanotubes)are used in many applications like FED, Micro wave amplifier, X ray source, etc. because of its suitable electrical, chemical and physical properties. Specially, CNTs are well used electron emitters for x ray source. Conventionally, thermionic type of tungsten filament x ray tube is widely employed in the field of bio medical and industrial application fields. However, intrinsic problems such as, poor emission efficiency and low imaging resolution cause the limitation of use of the x ray tube. To fulfill the current market requirement specifically for medical diagnostic field, we have developed rather a portable and compact CNT based x ray source in which high imaging resolution is provided. Electron sources used in X ray tubes should be well focused to the anode target for generation of high quality x ray. In this study, Pierce type x ray generation module was tested based its simulation results using by OPERA 3D code. Pierce type module is composed of cone type electrical lens with its number of them and inner angles of them that shows different results with these parameters. And some preliminary images obtained using the CNT x ray source were obtained. The represented images are the finger bone and teeth in human body. It is clear that the trabeculation shape is observed in finger bone. To obtain the finger bone image, tube currents of 250A at 42kV tube voltage was applied. The human tooth image, however, is somewhat unclear because the supplied voltage to the tube was limited to max. 50kV in the system developed. It should be noted that normally 60∼70kV of tube voltage is supplied in dental imaging. Considering these it should be emphasized that if the tube voltage is over 60kV then clearer image is possible. In this paper, we are discussed comparing between these experiment results and

  5. X rays from radio binaries

    International Nuclear Information System (INIS)

    Apparao, K.M.V.

    1977-01-01

    Reference is made to the radio binary systems CC Cas, AR Lac, β Per (Algol), β Lyr, b Per and Cyg X-1. It is stated that a thermal interpretation of the radiation from Algol requires a much larger x-ray flux than the observed value of 3.8 x 10 -11 erg/cm 2 /sec/keV in the 2 to 6 keV energy range. Observations of some non-thermal flares, together with the small size of the radio source in Algol, indicate that the radio emission is non-thermal in nature. The radio emission is interpreted as synchrotron radiation and it is suggested that the observed x-ray emission is due to inverse Compton scattering of the light of the primary star by the radio electrons. The x-ray emission from other radio binaries is also calculated using this model. The energy for the radio electrons can arise from annihilation of magnetic lines connecting the binary stars, twisted by the rotation of the stars. (U.K.)

  6. X-ray area monitor

    International Nuclear Information System (INIS)

    Nintrakit, N.

    1983-01-01

    The X-ray area monitor is a nuclear electronic device that is essential in radiation protection in high radiation laboratories, e.g. in medical diagnosis using X-rays and in industrial X-radiography. Accidentally the level of X-radiator may arise above the safe permissible level and in such a case the alarm system of the area monitor will work and disconnect the ac power supply form the X-ray unit. Principally the device is a radiation counter using G.M.tube as radiation detector with high voltage supply variable form 200 to 2,000 volts. The maximum count rate of the scaler is 1.5 MHz and the total count is displayed on 4 digit LED's. A time base is used to control the counting time, the frequency multiplier, radiation safety limit, comparator and the radiation hazard warning signal. The reliability of the instrument is further enhanced through the addition of the random correction circuit, and it is applicable both in X- and γ -radiation

  7. Russian medical X-ray engineering is confidently going into the year 2000

    International Nuclear Information System (INIS)

    Chikirdin, Eh.G.

    2000-01-01

    The achievements of the national medical X-ray technique are analyzed on the basis of the materials of the International exhibition Public health-99 which took place in November-December 1999, are analyzed. Great attention of the enterprises-developers was paid to fluorographic-digital equipment due to actuality of the lungs tuberculosis control in Russia. The characteristics for several fluorographic digital facilities, wherein by high technical characteristics it proves possible to improve the design and obtain low radiation doses for the patients and sufficiently low cost of the equipment, are presented. The characteristics of mobile facilities for X-ray diagnostics, including a fluorograph, mammograph, dental X-ray apparatus, are also presented. The attention was also paid to a X-ray diagnostic complex, feed devices, phantom sets, tomograph, dental apparatus and other developments in the area of X-ray technique. The outlooks for developments in the area of the national medical Xray technique are noted [ru

  8. Surveillance of X-ray machines in Israel

    International Nuclear Information System (INIS)

    Donagi, A.; Hai, J.; Kuszpet, M.

    1980-01-01

    A nationwide surveillance of X-ray machines is carried out in Israel by the Research Institute for Environmental Health, Ministry of Health. At present, diagnostic X-ray machines are surveyed at least once every two years, while dental machines are surveyed once every five years. The investigated parameters include measurement of output, scattered radiation, X-ray-light-field alignment, HVL, inherent filtration, structural shielding, etc. In order to compare X-ray techniques used in different hospitals in Israel, the NEXT (Nationwide Evaluation of X-ray Trends) program, which was developed by the BRH (Bureau of Radiological Health, USPHS) was utilized. On the basis of the findings of this project, necessary correction steps were taken in order to reduce the radiation doses to both personnel and patients. Further activities of the Institute include the estimation of doses delivered to pregnant women who were irradiated during the gestation period. This information is sent to a professional committee, which decides whether or not to perform an abortion. Recently, the new recommendations of ICRP 26 (International Commission on Radiological Protection) were implemented in Israel. Therefore, at present doses over 150 mR/month are reported to the Institute by the Soreq Personal Dosimetry Service, and the causes of this exposure are investigated. (author)

  9. Achromatic X-ray lenses

    International Nuclear Information System (INIS)

    Umbach, Marion

    2009-01-01

    This thesis presents first results on the development of achromatic refractive X-ray lenses which can be used for scientific experiments at synchrotron sources. First of all the different requirements for achromatic X-ray lenses have been worked out. There are different types of lenses, one type can be used for monochromatized sources when the energy is scanned while the spot size should be constant. The other type can be used at beamlines providing a broad energy band. By a combination of focusing and defocusing elements we have developed a lens system that strongly reduces the chromatic aberration of a refractive lens in a given energy range. The great challenge in the X-ray case - in contrast to the visible range - the complex refractive index, which is very similar for the possible materials in the X-ray spectrum. For precise studies a numerical code has been developed, which calculates the different rays on their way through the lenses to the detector plane via raytracing. In this numerical code the intensity distribution in the detector plane has been analyzed for a chromatic and the corresponding achromatic system. By optimization routines for the two different fields of applications specific parameter combinations were found. For the experimental verification an achromatic system has been developed, consisting of biconcave SU-8 lenses and biconvex Nickel Fresnel lenses. Their fabrication was based on the LIGA-process, including a further innovative development, namely the fabrication of two different materials on one wafer. In the experiment at the synchrotron source ANKA the energy was varied in a specific energy range in steps of 0.1 keV. The intensity distribution for the different energies was detected at a certain focal length. For the achromatic system a reduction of the chromatic aberration could be clearly shown. Achromatic refractive X-ray lenses, especially for the use at synchrotron sources, have not been developed so far. As a consequence of the

  10. 3D indoor modeling using a hand-held embedded system with multiple laser range scanners

    Science.gov (United States)

    Hu, Shaoxing; Wang, Duhu; Xu, Shike

    2016-10-01

    Accurate three-dimensional perception is a key technology for many engineering applications, including mobile mapping, obstacle detection and virtual reality. In this article, we present a hand-held embedded system designed for constructing 3D representation of structured indoor environments. Different from traditional vehicle-borne mobile mapping methods, the system presented here is capable of efficiently acquiring 3D data while an operator carrying the device traverses through the site. It consists of a simultaneous localization and mapping(SLAM) module, a 3D attitude estimate module and a point cloud processing module. The SLAM is based on a scan matching approach using a modern LIDAR system, and the 3D attitude estimate is generated by a navigation filter using inertial sensors. The hardware comprises three 2D time-flight laser range finders and an inertial measurement unit(IMU). All the sensors are rigidly mounted on a body frame. The algorithms are developed on the frame of robot operating system(ROS). The 3D model is constructed using the point cloud library(PCL). Multiple datasets have shown robust performance of the presented system in indoor scenarios.

  11. In-Season Yield Prediction of Cabbage with a Hand-Held Active Canopy Sensor.

    Science.gov (United States)

    Ji, Rongting; Min, Ju; Wang, Yuan; Cheng, Hu; Zhang, Hailin; Shi, Weiming

    2017-10-08

    Efficient and precise yield prediction is critical to optimize cabbage yields and guide fertilizer application. A two-year field experiment was conducted to establish a yield prediction model for cabbage by using the Greenseeker hand-held optical sensor. Two cabbage cultivars (Jianbao and Pingbao) were used and Jianbao cultivar was grown for 2 consecutive seasons but Pingbao was only grown in the second season. Four chemical nitrogen application rates were implemented: 0, 80, 140, and 200 kg·N·ha -1 . Normalized difference vegetation index (NDVI) was collected 20, 50, 70, 80, 90, 100, 110, 120, 130, and 140 days after transplanting (DAT). Pearson correlation analysis and regression analysis were performed to identify the relationship between the NDVI measurements and harvested yields of cabbage. NDVI measurements obtained at 110 DAT were significantly correlated to yield and explained 87-89% and 75-82% of the cabbage yield variation of Jianbao cultivar over the two-year experiment and 77-81% of the yield variability of Pingbao cultivar. Adjusting the yield prediction models with CGDD (cumulative growing degree days) could make remarkable improvement to the accuracy of the prediction model and increase the determination coefficient to 0.82, while the modification with DFP (days from transplanting when GDD > 0) values did not. The integrated exponential yield prediction equation was better than linear or quadratic functions and could accurately make in-season estimation of cabbage yields with different cultivars between years.

  12. Characterization of wood dust emission from hand-held woodworking machines.

    Science.gov (United States)

    Keller, F-X; Chata, F

    2018-01-01

    This article focuses on the prevention of exposure to wood dust when operating electrical hand-held sawing and sanding machines. A laboratory methodology was developed to measure the dust concentration around machines during operating processes. The main objective was to characterize circular saws and sanders, with the aim of classifying the different power tools tested in terms of dust emission (high dust emitter vs. low dust emitter). A test set-up was developed and is described and a measurement methodology was determined for each of the two operations studied. The robustness of the experimental results is discussed and shows good tendencies. The impact of air-flow extraction rate was assessed and the pressure loss of the system for each machine established. For the circular saws, three machines over the nine tested could be classified in the low dust emitter group. Their mean concentration values measured are between 0.64 and 0.98 mg/m 3 for the low dust emitter group and from 2.55 and 4.37 mg/m 3 for the high dust emitter group. From concentration measurements, a machine classification is possible-one for sanding machines and one for sawing machines-and a ratio from 1-7 is obtained when comparing the results. This classification will be helpful when a choice of high performance power tools, in terms of dust emission, must be made by professionals.

  13. Three-dimensional modeling of physiological tremor for hand-held surgical robotic instruments.

    Science.gov (United States)

    Tatinati, Sivanagaraja; Yan Naing Aye; Pual, Anand; Wei Tech Ang; Veluvolu, Kalyana C

    2016-08-01

    Hand-held robotic instruments are developed to compensate physiological tremor in real-time while augmenting the required precision and dexterity into normal microsurgical work-flow. The hardware (sensors and actuators) and software (causal linear filters) employed for tremor identification and filtering introduces time-varying unknown phase-delay that adversely affects the device performance. The current techniques that focus on three-dimensions (3D) tip position control involves modeling and canceling the tremor in 3-axes (x, y, and z axes) separately. Our analysis with the tremor data recorded from surgeons and novice subjects show that there exists significant correlation in tremor motion across the dimensions. Motivated by this, a new multi-dimensional modeling approach based on extreme learning machines (ELM) is proposed in this paper to correct the phase delay and to accurately model tremulous motion in three dimensions simultaneously. A study is conducted with tremor data recorded from the microsurgeons to analyze the suitability of proposed approach.

  14. Hand-held hyperspectral imager for chemical/biological and environmental applications

    Science.gov (United States)

    Hinnrichs, Michele; Piatek, Bob

    2004-03-01

    A small, hand held, battery operated imaging infrared spectrometer, Sherlock, has been developed by Pacific Advanced Technology and was field tested in early 2003. The Sherlock spectral imaging camera has been designed for remote gas leak detection, however, the architecture of the camera is versatile enough that it can be applied to numerous other applications such as homeland security, chemical/biological agent detection, medical and pharmaceutical applications as well as standard research and development. This paper describes the Sherlock camera, theory of operations, shows current applications and touches on potential future applications for the camera. The Sherlock has an embedded Power PC and performs real-time-image processing function in an embedded FPGA. The camera has a built in LCD display as well as output to a standard monitor, or NTSC display. It has several I/O ports, ethernet, firewire, RS232 and thus can be easily controlled from a remote location. In addition, software upgrades can be performed over the ethernet eliminating the need to send the camera back to the factory for a retrofit. Using the USB port a mouse and key board can be connected and the camera can be used in a laboratory environment as a stand alone imaging spectrometer.

  15. Hand-Held Photometer for Instant On-Spot Quantification of Nucleic Acids, Proteins, and Cells.

    Science.gov (United States)

    Li, Shi-Hao; Jain, Abhinav; Tscharntke, Timo; Arnold, Tobias; Trau, Dieter W

    2018-02-20

    This paper presents a novel hand-held photometer, termed "Photopette", for on-spot absorbance measurements of biochemical analytes. The Photopette is a multicomponent, highly portable device with an overall weight of 160 g, which fits within 202 mm × 47 mm × 42 mm. Designed in the form factor of a micropipette, Photopette integrates a photodiode detector with light emitting diodes (LEDs) to form a highly customizable photometer which supports a wide variety of applications within the wavelengths between 260 and 1050 nm. A dual-purpose disposable reflective tip was designed to act as a sample holder and a light-reflecting system, which is in stark contrast to the operation of mainstream spectrophotometers and photometers. Small volume analytes may be measured with low sample loss using this proprietary CuveTip. A user-friendly software application running on smart devices was developed to control and read the values from Photopette via a low-energy Bluetooth link. This one-step strategy allows measurements on-spot without sample transfer, minimizing cross-contamination and human error. The results reported in this paper demonstrate Photopette's great potential to quantify DNA, direct protein, and cell density directly within the laminar flow hood. Results are compared with a Nanodrop 2000c spectrophotometer, a mainstream spectrophotometer for small-volume measurements.

  16. Flexible CMOS low-noise amplifiers for beyond-3G wireless hand-held devices

    Science.gov (United States)

    Becerra-Alvarez, Edwin C.; Sandoval-Ibarra, Federico; de la Rosa, José M.

    2009-05-01

    This paper explores the use of reconfigurable Low-Noise Amplifiers (LNAs) for the implementation of CMOS Radio Frequency (RF) front-ends in the next generation of multi-standard wireless transceivers. Main circuit strategies reported so far for multi-standard LNAs are reviewed and a novel flexible LNA intended for Beyond-3G RF hand-held terminals is presented. The proposed LNA circuit consists of a two-stage topology that combines inductive-source degeneration with PMOS-varactor based tuning network and a programmable load to adapt its performance to different standard specifications without penalizing the circuit noise and with a reduced number of inductors as compared to previous reported reconfigurable LNAs. The circuit has been designed in a 90-nm CMOS technology to cope with the requirements of the GSM, WCDMA, Bluetooth and WLAN (IEEE 802.11b-g) standards. Simulation results, including technology and packaging parasitics, demonstrate correct operation of the circuit for all the standards under study, featuring NF13.3dB and IIP3>10.9dBm, over a 1.85GHz-2.4GHz band, with an adaptive power consumption between 17mW and 22mW from a 1-V supply voltage. Preliminary experimental measurements are included, showing a correct reconfiguration operation within the operation band.

  17. X-ray electromagnetic application technology

    International Nuclear Information System (INIS)

    2011-01-01

    The investigating committee aimed at research on electromagnetic fields in functional devices and X-ray fibers for efficient coherent X-ray generation and their material science, high-precision manufacturing, particularly for X-ray electromagnetic application technology from January 2006 to December 2008. In this report, we describe our research results, in particular, on the topics of synchrotron radiation and free-electron laser, Saga Synchrotron Project, X-ray waveguides and waveguide-based lens-less hard-X-ray imaging, X-ray nanofocusing for capillaries and zone plates, dispersion characteristics in photonics crystal consisting of periodic atoms for nanometer waveguides, electromagnetic characteristics of grid structures for scattering fields of nano-meter electromagnetic waves and X-rays, FDTD parallel computing of fundamental scattering and attenuation characteristics of X-ray for medical imaging diagnosis, orthogonal relations of electromagnetic fields including evanescent field in dispersive medium. (author)

  18. Aspergillosis - chest x-ray (image)

    Science.gov (United States)

    ... usually occurs in immunocompromised individuals. Here, a chest x-ray shows that the fungus has invaded the lung ... are usually seen as black areas on an x-ray. The cloudiness on the left side of this ...

  19. Soft x-ray Planetary Imager

    Data.gov (United States)

    National Aeronautics and Space Administration — The project is to prototype a soft X-ray Imager for planetary applications that has the sensitivity to observe solar system sources of soft  X-ray emission. A strong...

  20. Stabilized x-ray generator power supply

    International Nuclear Information System (INIS)

    Saha, Subimal; Purushotham, K.V.; Bose, S.K.

    1986-01-01

    X-ray diffraction and X-ray fluorescence analysis are very much adopted in laboratories to determine the type and structure of the constituent compounds in solid materials, chemical composition of materials, stress developed on metals etc. These experiments need X-ray beam of fixed intensity and wave length. This can only be achieved by X-ray generator having highly stabilized tube voltage and tube current. This paper describes how X-ray tube high voltage and electron beam current are stabilized. This paper also highlights generation of X-rays, diffractometry and X-ray fluorescence analysis and their wide applications. Principle of operation for stabilizing the X-ray tube voltage and current, different protection circuits adopted, special features of the mains H.V. transformer and H.T. tank are described in this report. (author)