WorldWideScience

Sample records for glass-rod thermoluminescent dosemeters

  1. Dosimertry characteristics and performance comparisons: Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Lee, J.H.; Lin, M.S.; Hsu, S.M.; Chen, I.J.; Chen, W.L.; Wang, C.F.

    2009-01-01

    This paper describes the dosimetry characteristic comparison of environmental radiophotoluminescent glass dosemeter (RPLGD) and thermoluminescent dosemeter (TLD) systems employed at the Institute of Nuclear Energy Research (INER, Taiwan). The luminescence centers of TLD disappeared by reading process, and repetition of measurement is impossible. RPLGDs can be repeatedly read and keep the luminescence centers for a long time. The RPLGD fading is about 1% within 30 days after being exposed to ionizing radiation. Environmental cumulative doses measured by a RPLGD and a TLD were compared at the same monitoring points of INER during two years. The sensitivity of TLD decreased with gradual loss due to fading at higher temperatures. The difference between results obtained by RPLGD and TLD is discussed mainly with respect to the dependence on the ambient temperature. In addition, this research also refers to the American National Standard Institute (ANSI) Draft Standard N13.29 (1996) to organize the environmental dosimetry performance tests of the INER's RPLGDs and the TLDs. The results mean that both kinds of dosemeters can meet the performance test criteria. An accreditation procedure for the institutes which provide the environmental dosimetry services in Taiwan was suggested based on the comparison results of these two dosimetry systems

  2. Selected references on thermoluminescent dosemeters for personnel monitoring

    International Nuclear Information System (INIS)

    Lawson, Helen

    1976-08-01

    This bibliography contains references on: specific thermoluminescent dosemeters used for personnel dosimetry; comparisons of thermoluminescent and other dosemeters for personnel monitoring and read-out apparatus for thermoluminescent dosemeters. An appendix contains selected foreign language references. (author)

  3. Performance tests on the NRPB thermoluminescent dosemeter

    CERN Document Server

    Shaw, K B

    1977-01-01

    Performance tests on the thermoluminescent dosemeter, designed at NRPB for use in the automated personal dosimetry system, are described. An ultra-thin lithium borate dosemeter has been developed for skin absorbed dose measurement. The X-ray, gamma-ray and beta-ray energy response of the dosemeter has been investigated and the angular response for the dosemeter has been examined. The annealing, read-out and stabilisation procedures for the dosemeter are described.

  4. Personnel monitoring of radiations with thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Miano, S.C.

    1987-01-01

    The basics of personnel dosimetry technics, used by the Radiologic Protetion and Assessorie Service (SAPRA) are presented, consisting on use of thermoluminescent and CaSO 4 :Dy monitors in aggregated pellets by Teflon. The characteristics of this dosemeters, relating to the sensitivity, energetic dependence, spike temperature, characteristic emission curve, decay and light effect are shown. The thermoluminescent dosemeter measure system and the personnel monitoring system are also described. (C.G.C.) [pt

  5. Needle-type differential thermoluminescent dosemeter

    International Nuclear Information System (INIS)

    Kubu, M.

    1988-01-01

    The dosemeter serves to measure the depth distribution of absorbed dose in a phantom in mixed neutron and gamma radiation fields. It consists of a stratified cylindrical absorber in which alternate moderator and cadmium layers. Inside are two bores along the longitudinal axis for tubes filled with 6 LiF and 7 LiF powder thermoluminescent dosemeters. The device is easily portable and allows to carry out measurements with a high level of reproducibility. (J.B.). 4 figs

  6. Personal radiation monitoring with thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Miano, S.C.

    1987-01-01

    The technique of personal dosimetry used by SAPRA (Servico de Assessoria e Protecao Radiologica S/C Ltda., Brazil) is presented. Thermoluminescent monitors and CaSO 4 : Dy are used in pastilles united by teflon. Characteristics of the dosemeters are briefly reported. The system of thermoluminescent measurement, designed and constructed by SAPRA, and the system of personal monitoring are described. (M.A.C.) [pt

  7. Thermoluminescence sensitivity variations in LiF PTFE dosemeters incurred by improper handling procedures

    CERN Document Server

    Mason, E W; MacKinlay, Alistair F; Saunders, D

    1975-01-01

    A systematic study of some anomalous darkening effects and thermoluminescence sensitivity variations observed in LiF:PTFE thermoluminescent dosemeters is described. Various likely causes of such effects have been investigated. The manufacturer's recommended cleaning procedures have been found to be inadequate and, in some cases, have been found to actually promote discolouration of the dosemeters. Recommendations are given for the successful use of LiF:PTFE thermoluminescent dosemeters in personal dosimetry.

  8. Thermoluminescence dosemeter for personal dose equivalent assessment

    International Nuclear Information System (INIS)

    Silva, T.A. da; Rosa, L.A.R. da; Campos, L.L.

    1995-01-01

    The possibility was investigated of utilising a Brazilian thermoluminescence individual dosemeter, usually calibrated in terms of photon dose equivalent, for the assessment of the personal dose equivalent, H p (d), at depths of 0.07 and 10 mm. The dosemeter uses four CaSO 4 :Dy thermoluminescent detectors, between different filters, as the sensitive materials. It was calibrated in gamma and X radiation fields in the energy range from 17 to 1250 keV. Linear combinations of the responses of three detectors, in this energy range, allow the evaluation of H p (0.07) and H p (10), for radiation incidence angles varying from 0 to 60 degrees, with an accuracy better than 35%. The method is not applicable to mixed photon-beta fields. (author)

  9. Environmental radiation measurements using lithium fluoride thermoluminescence dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Driscoll, C.M.H.; Green, B.M.R.; McKinlay, A.F.; Richards, D.J. (National Radiological Protection Board, Harwell (UK))

    1984-01-01

    The National Radiological Protection Board is involved in a large scale environmental survey of radiation levels in homes throughout the United Kingdom. Passive radon and gamma ray dosemeters are posted to a representative sample of households. Lithium fluoride thermoluminescence dosemeters are used to assess natural gamma radiation and are left in the measurement location for a period of six months before being returned to the Board for processing. As a preliminary to the national survey, the Board has been engaged on several limited surveys in regions of igneous and sedimentary geology. Experience gained in these limited surveys and from standardisation studies using environmental dosemeters are reported.

  10. Characterization of the personal thermoluminescent dosemeter of LiF: Mg, Ti + Ptfe

    International Nuclear Information System (INIS)

    Azorin N, J.; Gutierrez C, A.; Gonzalez M, P.

    1991-01-01

    The objective of this work is to characterize the thermoluminescent dosemeters taken place in the laboratory in form of pellets of LiF: Mg, Ti + Ptfe like personal dosemeters, subjecting them to the operation tests proposed by the international standards and comparing them with the TLD-100, the Tl dosemeter more used at the moment for personal dosimetry

  11. Development trends of radiophotoluminescent glass dosemeter

    International Nuclear Information System (INIS)

    Ikegami, T.

    2004-01-01

    RPL glass dosemeter has been recently recognized to have a good performance as accumulation dosemeter and has been routinely used for personal dosimetry and environmental radiation monitoring. Furthermore, its applicable field is being extending to medical radiation measurement. The history of RPL glass dosemeter is very long. It was born in USA in 1950s and after that it was improved in Japan. And it was used as personal dosemeter in 1970s. But, in those days, RPL glass dosemeter was not suitable to low dose measurement due to some handling problems. So, its use had been reduced gradually. The author has broken through any past problems, by mainly realizing the pulsed UV excitation method. In this paper, the development history, principle and features, and the development good results of the pulsed UV excitation method are summarized, including the introduction of recent RPL glass dosimetry products. (author)

  12. The NRPB automated thermoluminescent dosemeter and dose record keeping system

    International Nuclear Information System (INIS)

    Dennis, J.A.; Marshall, T.O.; Shaw, K.B.

    1975-03-01

    This report describes the thermoluminescent personal radiation dosemeter and its associated automated processing equipment, which are being developed by the National Radiological Protection Board, together with the operation of a computerised dosemeter issue and record keeping system. The main justifications for introducing these systems are improvements in the organizational efficiency of the maintenance of individual dose records, a more flexible and accurate dosimetry system, and economics in operational costs. The dosemeter is based on a numbered aluminium plate containing two lithium fluoride in polytetrafluorethylene disks for the measurement of surface and body dose. This dosemeter is wrapped in thin plastic and labelled with the wearer's name and address. On return, the dosemeter is checked automatically for radioactive contamination; it is unwrapped and evaluated; the dose readings are included in the wearer's stored dose record; the dosemeter is annealed and is then available for re-issue to another wearer. Dose reports and warnings are automatically issued to the wearer or his employer. (author)

  13. Dosimetric of extremities with Dosemeters thermoluminescent in Cuba

    International Nuclear Information System (INIS)

    Molina Perez, D.; Diaz Bernal, E.; Vera Alonso, L.

    1998-01-01

    From final of the year 1995 in the CPHR implement the service of monitoring of the extremities using Dosemeter thermoluminescent (TL). The dosemeter consists on a metallic ring with a circular hole where a detector of LiF:Mg,Ti is placed (model JR1152C) of 5x5x0.9 mm 3 , covered by a fine layer of polyethylene. In the work the characteristic dosimetric as of the dosemeter is studied it satisfies the main requirements for their use in the monitoring from the exhibition to radiation photonic of the extremities. The doses are also presented registered during the first two years of operation of the service. The results obtained until the moment point out to you practice them of nuclear medicine, radiotherapy and production of substances radioactive how as of more contribution

  14. Solid thermoluminescent dosemeter of sodium tetraborate and brazilian fluorite sensible to thermic neutrons

    International Nuclear Information System (INIS)

    Fratin, L.; Cruz, M.T. da

    1987-01-01

    The production of solid thermoluminescent dosemeters sensible to thermic neutrons is described. The stages of the production are described: vitrification of sodium tetraborate, mixture and cold pressing, syntherization. The obtention of a CaF 2 : natural dosemeter with NaCl which presents sensibility to gamma radiation similar to CaF 2 : natural dosemeter with Na 2 B 4 O 7 is studied. (M.A.C.) [pt

  15. Thermoluminescent dosemeter in a X-ray diffractometer

    International Nuclear Information System (INIS)

    Mendoza A, D.; Gonzalez M, P.; Falcon B, T.; Castano, V.M.

    1999-01-01

    In this work it was presented the results obtained of the dosimetry which was realized in a X-ray diffractometer for powders, trademark Siemens D5000, using the thermoluminescent signal generated by the X-rays in the commercial dosemeter TLD-100 of Harshaw, US. In according to the results obtained, the radiation quantity received by an analysed material in the diffractometer, will be proportional to exposure time and it can vary from unities until tenths of grays. These results are very outstanding when are analysed crystalline materials in a diffractometer, for knowing the present crystalline phases, mainly if these are highly sensitive to the ionizing radiation, as it is the case of the thermoluminescent materials. (Author)

  16. About the measurements systems with pen and thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Cortes I, M.E.; Ramirez G, F.P.

    1998-01-01

    In this work it is presented dosimetric data obtained with pen and thermoluminescent dosemeters, which are used by the Occupational Exposure Personnel (OEP) of the Mexican Petroleum Institute (IMP)(1). It was marked several great characteristics as for example, the differences among units which use one and another dosemeter type. Likewise, it is given to know diverse problems that were had in the IMP at relating the data obtained with these dosemeters (which utilizes OEP) and the ICRP 60 recommendations 1990. One of the most important difficulties is to satisfy the recommended limits by ICRP, particularly those that are referring to the units and their complex calculations. With respect to the unities, the ICRP makes reference at the concepts 'dose equivalent' and 'effective dose' with the sievert unit, that the General Regulations for Radiological Safety associates with 'dose equivalent' and 'effective dose equivalent'. It was illustrated the type of dosimetric statistics which are obtained with the TLD lectures and a OEP pen dosemeter during 1997. (Author)

  17. Automated read-out of thermoluminescence dosemeters in a centralized individual monitoring service

    International Nuclear Information System (INIS)

    Toivonen, M.

    The organizational problems in maintaining centralized individual monitoring service with erasable and re-usable dosemeters are evaluated. Design criteria for an automated thermoluminescence reader are laid down. It is characteristic for the planning of the monitoring system that the issuing of dosemeters can be arranged without having two dosemeters for each worker. A home made reader designed to fullfil these criteria is presented. The use of a standard barcode and a standard optical barcode reader in identification of dosemeters is described. A method of using a minicomputer in preparing the self-fastening identification labels, in printing mailing lists and in printing results is described

  18. Measurement of gamma radiation doses at the RA reactor by thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Prokic, M.

    1974-01-01

    This paper presents the procedures and gamma radiation doses measured at the exit from the horizontal experimental channel HK-5, vertical experimental channel VK-5 and in the thermal column of the RA reactor in Vinca. Measurement of gamma radiation dose in the mixed intense gamma and neutron radiation field was done by two types of thermoluminescent dosemeters, LiF (TLD-700) and CaF 2 (TLD-08). Gamma dose in the VK-5 was measured in the air and on the bottle filled with tissue-equivalent solution. Increase of the dose on the surface of the bottle was 2.3 compared to the gamma dose value in the air. Correction for the influence of neutrons having different energies was done by using the known sensitivity values of both TL dosemeter types for thermal, intermediate and fast neutrons. Results showed that the TLD-700 dosemeter contains 5 time more Li-6 isotopes (0.035%) than the declared value causing increased neutron sensitivity of this dosemeter. This paper includes numerical sensitivity data for neutrons of different energies for both types of TL dosemeters. Neutron sensitivity values for TLD-700 are related to LiF with 0.035% of Li-6 isotope. Result of measurement have also shown that the CaF 2 :Mn (TLD-08) thermoluminescent dosemeter is more suitable for gamma radiation dose measurements in mixed n-gamma fields with intensive neutron fluxes due to lower neutron sensitivity compared to TLD-700 [sr

  19. Utilization of thermoluminescent dosemeters for determination of exposure or absorbed dose in a radiation gamma or X radiation field with unknown spectral distribution

    International Nuclear Information System (INIS)

    Rosa, L.A.R. da.

    1981-06-01

    Having in view the choice of the best pair of dosemeters to be used in the 'Tandem' method, the main response characteristics of LiF:Mg, Ti, Li 2 B 4 O 7 :Mn, CaSO 4 Dy, CaF 2 :Mn and CaF 2 :Dy thermoluminescent dosemeters and also some critical parameters in their calibration and evaluation processes were studied. Three different physical forms of TLD's were investigated: hot pressed chips, disc teflon dosemeters and glass mini TLD's. Their calibration factors were obtained for the energy of Cobalt-60 gamma rays. Their energy dependences normalized to 60 Co radiation were determined using spectral width as parameter. 'Tandens' formed by all TLD's evaluated were compaired. (E.G.) [pt

  20. CaSO4: Dy + Teflon thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Campos, L.L.

    1986-01-01

    A pellet dosemeter of CaSO 4 : Dy + Teflon was developed at IPEN. CaSO 4 : Dy thermoluminescent phosphor, grown in the Dosimetric Materials Production Laboratory was chosen, due to its high sensitivity, ease of preparation and comparatively low cost. Pellets were produced by cold pressing and sintering a mixture of CaSO 4 : Dy and Teflon powders. Extensive work was done to study in detail all CaSO 4 : Dy pellets characteristics from the point of view of dosimetry with the purpose of introducing it in the routine use. A filter combination providing an energy independent response from 20 KeV to 1,25 MeV was obtained. The dosemeter consists of three pellets sealed between two thin plastic sheets and placed under plastic and lead filters. The combination of these tree filters allows the exposure as well as the energy determination of an unknown source. (Author) [pt

  1. One year of a thermoluminescent dosemeter and dose record keeping service

    International Nuclear Information System (INIS)

    Dennis, J.A.

    1979-01-01

    A brief description is given of the National Radiological Protection Board's (NRPB) thermoluminescent dosemeter. The automated processing of these dosemeters is linked to a computer-based system of dose records. The postal system is currently used by about 10% of the establishments for which the NRPB provides a personnel monitoring service, but it will soon be extended to other establishments. The distributions of individual whole body doses in 1977 are tabulated. The significant advantage of this new system lies in the improvements in accuracy, completeness, and convenience to the employers of record keeping. (U.K.)

  2. Glass as a gamma ray dosemeter

    International Nuclear Information System (INIS)

    Sutrisno Puspodikoro.

    1978-01-01

    The advantages of glass as a γ-rays dosemeter are studied. Experiments have shown that ordinary microscope object glass can be used as a dosemeter, which dose range for linear response extends from about 10 4 -10 6 rads. Heat treatment of the irradiated samples accelerates the initial fading of coloration and stabilizes the residual optical density. On the other side cooling of them retards the initial fading. (author)

  3. Solid thermoluminescent dosemeter of sodium tetraborate and brazilian fluorite sensible to thermal neutrons

    International Nuclear Information System (INIS)

    Fratin, L.; Cruz, M.T. da

    1987-01-01

    A solid termoluminescent dosemeter of sodium tetraborate and brazilian fluorite sensible to thermal neutrons is described. The nuclears reactions 1) 10 B + n → 7 Li + He + Q1 (6,1%) where: Q1=2,79 MeV and Eα1 = 1,758 MeV and 2) 10 B + n → 7 Li* + 4 He + Q2 (93,9%) where: Q2 = 2,316 MeV and E2α 2 = 1,474 MeV are responsible by the thermoluminescent response of the thermal neutrons dosemeters. The stages in the fabrication process of this dosemeter of which are:1) sodium tetraborate vitrification, 2) mixture and pressing 3) sintering are cited. The obtainment of a natural fluorite dosemeter with sodium chloride is also shown. (C.G.C.) [pt

  4. Determination of the dose rapidity of a 90 Sr beta radiation source using thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Gonzalez M, P.R.; Azorin N, J.; Rivera M, T.

    2000-01-01

    The thermoluminescent dosemeters developed in Mexico, have been used efficiently in environmental and personal dosimetry. When the dose rate of some source is not known can be estimated with the use of thermoluminescent dosemeters taking in account the geometrical array used in the irradiations for reproducibility of the results in posterior irradiations. In this work it was estimated the dose rate of a 90 Sr- 90 Y beta radiation source which is property of the Nuclear Sciences Institute, UNAM, therefore it was l ended to the Metropolitan Autonomous University- Iztapalapa Unit for the characterization of new Tl materials, taking account of the institutional collaboration agreements. (Author)

  5. Method of making isodose curve using thermoluminescence dosemeter

    International Nuclear Information System (INIS)

    Bakar Ramain, A.

    1975-01-01

    Line source of 60 Co in the form of needles and tubes are extensively used in radiotherapy in moulds, implants, and intracavitary techniques for the treatment of malignant lesions. It is important to have isodose distributions in tissue for those source for purposes of treatment planning. The isodose distributions have been obtained experimentally by using tinny lithium-fluoride (Lsub(i)F) thermoluminescent dosemeters (TLD) and they are compared to the theoretical results. The distinct advantages of Lsub(i)F (TLD) in such measurements are briefly discussed. (author)

  6. Dose profile measurement in computerized axial tomography equipment using thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Azorin V, J.C.; Falcony, C.; Azorin N, J.

    2000-01-01

    In this work are presented the results about measuring the radiation dose profile in two equipment of computerized axial tomography (Tac). Thermoluminescent dosemeters (Dtl) of LiF, Mg, Cu, P + Ptfe in form of disks were used which were developed and made in Mexico. The results showed that Dtl are appropriated for these type of studies. (Author)

  7. Thermoluminescent response of dosemeters TLD-100 exposed to distinct baker protocols, irradiation and reading

    International Nuclear Information System (INIS)

    Estrada E, D.

    2002-01-01

    The interest that motivated the realization of this work is the opposing discrepancy among the values of the efficiency thermoluminescent of heavy charged particles (PCP) regarding gamma radiation (η PCP , γ) reported in different works and gathered recently for Horowitz (Ho 01). The measures of relative thermoluminescent efficiency have importance in relation to the dosimetry associated to medical therapy with heavy charged particle. The measurement of η PCP , γ depends of multiple experimental factors. To understand this dependence it was quantified the effect of using different procedures experimental in the estimate of the relative thermoluminescent efficiency of dosemeters TLD-100 (LiF: Ti; Mg) for protons of 3 MeV regarding gamma rays of 60 Co (η PCP , γ), varying the lot, the presentation, the one baked and the team reader. When was used the same global procedure values they were obtained for (η PCP , γ) inside the interval [0.30, 0.58] for the peak 5 and [0.35, 0.76] for the total thermoluminescent signal (Tl). At the use different equipment, baked and lot are obtained maximum differences of 30%, 12% and 6% respectively in the values of (η PCP , γ) for the peak 5, and 25%, 28% and 7% for the total Tl signal. The changes more significant of 36% and 44% for peak 5 and total signal respectively are obtained when changing the presentation of the dosemeter. As a complementary investigation it was studied the distribution of the Tl signal of each peak in the deconvolution of the curve of brightness with regard to the total signal in exposed dosemeters to high fluence of protons (1x10 10 p/cm 2 ) varying the experimental protocol. It was found that the distribution of the Tl signal has a difference of the order of 5% when changing the used reading team. It was observed that the contribution of the peak 5 regarding the total signal are preserved when changing baked and that it results to be 8% greater for the presentation in dosemeters of 1 x 1 x 1 mm 3

  8. Development and testing of a thermoluminescent dosemeter for mixed neutron-photon-beta radiation fields

    International Nuclear Information System (INIS)

    Zummo, J.J.; Liu, J.C.

    1998-08-01

    A new four-element thermoluminescent (TL) dosemeter and dose evaluation algorithm have been developed and tested to better characterize personnel exposure in mixed neutron-photon-beta radiation fields. The prototype dosemeter is based on a commercially available TL card (with three LiF-7 chips and one LiF-6 chip) and modified filtration elements. The new algorithm takes advantage of the high temperature peak characteristics of the LiF-6 element to better quantify the neutron dose component. The dosemeter was tested in various radiation fields, consisting of mixtures of two radiation types typically used for dosemeter performance testing, as well as mixtures of three radiation types to simulate possible exposure conditions. The new dosemeter gave superior performance, based on the tolerance levels, when using the new algorithm as compared to a conventional algorithm that did not use the high temperature peak methodology. The limitations and further improvements are discussed

  9. Characterization of a two-component thermoluminescent albedo dosemeter according to ISO 21909

    Energy Technology Data Exchange (ETDEWEB)

    Martins, M.M., E-mail: marcelo@ird.gov.b [Instituto de Radioprotecao e Dosimetria (IRD/CNEN), Av. Salvador Allende s/n, CEP 22780-160, Rio de Janeiro, RJ (Brazil); Mauricio, C.L.P., E-mail: claudia@ird.gov.b [Instituto de Radioprotecao e Dosimetria (IRD/CNEN), Av. Salvador Allende s/n, CEP 22780-160, Rio de Janeiro, RJ (Brazil); Pereira, W.W., E-mail: walsan@ird.gov.b [Instituto de Radioprotecao e Dosimetria (IRD/CNEN), Av. Salvador Allende s/n, CEP 22780-160, Rio de Janeiro, RJ (Brazil); Silva, A.X. da, E-mail: ademir@con.ufrj.b [Coordenacao dos Programas de Pos-Graduacao em Engenharia, COPPE/PEN Caixa Postal 68509, CEP 21941-972, Rio de Janeiro, RJ (Brazil)

    2011-05-15

    A two-component thermoluminescent albedo neutron monitoring system was developed at Instituto de Radioprotecao e Dosimetria, Brazil. As there is no Brazilian regulation for neutron individual monitoring service, the system was tested according to the ISO 21909 standard. This standard provides performance and test requirements for determining the acceptability of personal neutron dosemeters to be used for the measurement of personal dose equivalent, H{sub p}(10), in neutron fields with energies ranging from thermal to 20 MeV. Up to 40 dosemeters were used in order to accomplish satisfactorily the requirements of some tests. Despite operational difficulties, this albedo system passed all ISO 21909 performance requirements. The results and problems throughout this characterization are discussed in this paper.

  10. Determination of the dose index in computerized tomography using thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Azorin, J.C.; Calderon, A.; Azorin, J.

    2004-01-01

    In this work the obtained results of the determination of the dose index are presented in thorax studies in computed tomography and helical tomography carried out in Mexico using thermoluminescent dosemeters of LiF: Mg,Cu,P + Ptfe developed and manufactured in our country. The results showed that under similar conditions of irradiation and operation (pitch = 1), significant differences don't exist among the doses absorbed measures in the phantom due to the two types of used tomographs. (Author)

  11. Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study

    NARCIS (Netherlands)

    Koivisto, J.; Schulze, D.; Wolff, J.E.H.; Rottke, D.

    2014-01-01

    Objectives: The objective of this study was to compare the performance of metal oxide semiconductor field-effect transistor (MOSFET) technology dosemeters with thermoluminescent dosemeters (TLDs) (TLD 100; Thermo Fisher Scientific, Waltham, MA) in the maxillofacial area. Methods: Organ and effective

  12. The thermoluminescence of Toshiba FD-1 and FD-7 RPL glass

    International Nuclear Information System (INIS)

    Croft, S.; Weaver, D.R.; Matthews, R.

    1996-01-01

    The thermoluminescent response of Toshiba FD-1 and Toshiba FD-7 radiophotoluminescent dosimetry glass to 60 Co γ radiation has been studied at doses of up to 66 kGy air kerma and 590 kGy air kerma respectively. The FD-1 glass was in the form of rods 1 mm in diameter by 6 mm long whereas the FD-7 glass was in the form of blocks of 4.6 mm square and 1.47 mm thick. The FD-1 rods had a pre-dose signal equivalent to 180 Gy and were useable up to the maximum dose studied although the response-dose curve was slightly supralinear at low doses. The FD-7 samples had a predose of 6.2 Gy and began to show severe saturation beyond about 1 kGy. The specific TL efficiencies (counts. mg. -1 .Gy -1 ) of the two glass types have been compared with LiF and BeO and found to be some six orders of magnitude lower. (author)

  13. Performance tests of a CaSO4:Dy based thermoluminescence personal dosemeter

    International Nuclear Information System (INIS)

    Azorin, J.; Gutierrez, A.; Niewiadomski, T.

    1986-01-01

    In order to replace LiF commercial thermoluminescence dosemeters (TLDs) now used in personnel monitoring, we developed a TLD personnel monitoring badge based on pellets of CaSO 4 :Dy bound in pure KBr placed in plastic holders provided with 0.8 mm Pb filters to compensate for the energy dependence of CaSO 4 :Dy. Performance tests, based on the proposals of international standards for TLDs intended to be used as general deep tissue personal dosemeters, were done by developing the detailed test methods both for the tests required by the International Atomic Energy Agency (fading, linearity, effect of fluorescent light and sunlight and stability under standard climatic conditions) and to meet the requirements of the International Electrotechnical Commission (batch homogeneity, detection threshold, stability of zero point repeatibility and residual). Initial tests showed a high residual and an increase of inhomogeneity. However, these drawbacks were overcome by changing the annealing procedure and the readout conditions. Using these new conditions TLDs met all the requirements for personal dosemeters. (author)

  14. Thermoluminescent and optical processes in alkaline halogenides dosemeters contaminated with Europium

    International Nuclear Information System (INIS)

    Barboza F, M.; Melendrez, R.; Castaneda, B.; Pedroza M, M.; Chernov, V.; Perez S, R.; Aceves, R.

    2000-01-01

    Recent research results are presented about the properties of the optical processes of photo transferred thermoluminescence (TLFT), optical whitening (BO), thermoluminescence induced by light (TLL) and its effect in the thermoluminescent curve (Tl) produced by ionizing and non-ionizing radiation. The systematic analysis of all these processes, acquires a singular importance due that actually the alkaline halogenide crystals are object of intense investigations which analyse their potential applications as detectors and radiation dosemeters through stimulated optical luminescence techniques or thermoluminescence. The obtained data show that the Tl curve of material with this nature can be enormously affected by exposure of phosphorus to the environmental light or UV. This is in part due to liberation processes of charge bearers are shouted and makes a subsequent trapping in less temperature traps; at the same time that induce changes in the intensity of determined Tl bands. Additionally, also it is observed that mentioned phenomena are related as with wavelength of incident light as of the illumination time. Finally, the obtained information allows to conclude that although the illumination effect is extremely complex, it is associated and can be explained mainly with phenomena that implicate the electrons excitation trapped in form of F centers and trapping mechanisms or radioactive and non-radioactive recombination. (Author)

  15. Photon and fast neutron dosimetry using aluminium oxide thermoluminescence dosemeters in a pool-type research reactor

    International Nuclear Information System (INIS)

    Santos, J.P.; Marques, J.G.; Fernandes, A.C.; Osvay, M.

    2007-01-01

    Al 2 O 3 :Mg,Y thermoluminescence (TL) dosemeters were used to measure photon and fast neutron doses in the mixed radiation field of the Portuguese Research Reactor. The dosemeters were irradiated in core positions under a photon dose rate of the order of 10 4 Gy/h and a fast neutron flux in the range of 10 9 -10 11 n/cm 2 /s. In order to evaluate the ability of the TL dosemeters for mixed field dosimetry at the research reactor, the measurements were compared with results obtained via conventional methods. The agreement between the different methods is better than 13% for the determination of photon doses and within 5% for the determination of neutron fluxes in mixed fields

  16. Technical specification of the NRPB thermoluminescent dosemeter used for the measurement of body dose and skin dose

    CERN Document Server

    Shaw, K B

    1977-01-01

    This report specifies the NRPB thermoluminescent dosemeter used for the measurement of radiation dose in tissue at a depth of 700 mg cm sup - sup 2 (body dose) and at a depth of 5-10 mg cm sup - sup 2 (skin dose).

  17. Comparison of responses of thermoluminescent dosemeters irradiated by soft x-rays at very low and very high dose rate levels

    International Nuclear Information System (INIS)

    Pietrikova-Farnikova, M.; Krasa, J.; Juha, L.

    1994-01-01

    Recent great progress in construction and application of bright sources of soft X-rays gave a strong impetus for the development of methods of their dosimetric diagnostics. The soft X-ray sources are primarily represented by synchrotron radiation sources and by sources based on laser-produced plasma, including X-ray lasers. Their characteristics spread over a very wide region of photon energies, peak and average powers and densities. From our preliminary experiments it follows that thermoluminescent dosemeters can serve as a suitable tool for the determination of these characteristics. Problem lies in the fact that routine use of the thermoluminescent dosemeters for the dosimetry of soft X-rays requires their spectral calibration, which can be carried out with low peak power sources (synchrotron radiation and radionuclide sources). On the contrary, many important sources, especially these based on laser-produced plasmas, exhibit a very high peak power, i.e. dosemeters are irradiated at extremely high dose rate. In comparative experiments carried out with laser-produced plasmas and radionuclides using TLD 200 (CaF 2 :Dy) and GR 200A (LiF:Mg,Cu,P) it was satisfactorily proven that total thermoluminescent signals are independent of the dose rate. Dependence of glow curve shapes on the dose, dose rate and photon energy were equally determined

  18. Calibration of thermoluminescence skin dosemeter response to beta emitters found in Ontario Hydro nuclear power stations

    International Nuclear Information System (INIS)

    Walsh, M.L.; Agnew, D.A.; Donnelly, K.E.

    1984-01-01

    The response of the Ontario Hydro Thermoluminescence Dosimetry System to beta radiation in nuclear power station environments was evaluated. Synthetic beta spectra were constructed, based on activity samples from heat transport systems and fuelling machine contamination smears at nuclear power stations. Using these spectra and dosemeter energy response functions, an overall response factor for the skin dosemeter relative to skin dose at 7 mg.cm -2 was calculated. This calculation was done assuming three specific geometries: (1) an infinite uniformly contaminated plane source at a distance of 33 cm (50 mg.cm -2 total shielding) from the receptor; (2) an infinite cloud surrounding the receptor; (3) a point source at 33 cm. Based on these calculations, a conservative response factor of 0.7 has been chosen. This provides an equation for skin dose assignment, i.e. Skin Dose = 1.4 x Skin Dosemeter Reading when the skin dosemeter is directly calibrated in mGy(gamma). (author)

  19. Thermoluminescent dosemeters characterization for patient dosimetry in diagnostic radiology preliminary results

    International Nuclear Information System (INIS)

    Castro, William J.; Squair, Peterson L.; Gonzaga, Natalia B.; Nogueira, Maria S.; Silva, Teogenes A. da

    2009-01-01

    The determination of the metrological characteristics of thermoluminescent (TL) dosimeters plays an important role in dosimetry of patients submitted to x-ray examinations for diagnostic purpose. Entrance surface doses can be measured with TL dosimeters to verify the compliance with the diagnostic reference levels. Organ doses can be estimated through TL measurements in an anthropomorphic phantom which it allows the radiation risk assessment. In this work, LiF:Mg,Ti (TLD-100) rod and chip thermoluminescent dosimeters were characterized considering their use for patient dosimetry in computerized tomography and mammography. Preliminary results showed that TL dosimeters have a response reproducibility of 7.8% and 4.8% and homogeneity of 18.4% and 6.5% for rod and chip shapes, respectively. (author)

  20. Determination of the parameters of traps in thermoluminescent dosemeters of ZrO2: Eu + Ptfe

    International Nuclear Information System (INIS)

    Nieto H, B.; Azorin N, J.; Rivera M, T.

    2002-01-01

    In this work the manufacture of dosemeters in pellets form of ZrO 2 : Eu + Ptfe is described; which after were exposed to UV radiation (260 nm) were generated the corresponding thermoluminescent curves and in the basis to the properties of symmetry of the curves it was determined the kinetic order using the Chen and the Balarin criteria. (Author)

  1. A pocket type thermoluminescent personnel dosimeter

    International Nuclear Information System (INIS)

    Vora, K.G.; Nagpal, J.S.; Pendurkar, H.K.; Gangadharan, P.

    1979-01-01

    A pocket type thermoluminescent personnel dosemeter using CaSO 4 : Dy phosphor is described. Two glass capillaries containing the phosphor are fitted into a plastic tube and covered by a cylindrical filter. The combination is fitted into an ink barrel of a fountain pen. The response of this Dy glass dosimeter was studied for various incident photon energies. A uniform response over the energy range from 33 keV to 1250 keV is achieved. (A.K.)

  2. Al2 O3:Cr,Ni: a possible thermoluminescent dosemeter

    International Nuclear Information System (INIS)

    Mariani R, Francisco; Roman B, Alvaro; Saavedra S, Renato; Ibarra S, Angel

    1996-01-01

    Results from a study on the thermoluminescent (Tl) emission from Al 2 O 3 :Cr,Ni are presented. The measurements were obtained for evaluation of the Al 2 O 3 :Cr,Ni dosimetric properties. Different crystal batches were exposed to two kind of ionizing radiation (X-ray and β - ). The Tl spectrum has a main peak with high thermal and optical stability, deviating from linearity for doses lower than 3.6 Gy. Furthermore, this material shows advantages (thermal resistance, reusability, multiple heating cycles) compared to TLD-100. Measured Al 2 O 3 :Cr,Ni properties indicate that it could be used as a dosemeter. (author)

  3. Thermoluminescence and applications

    International Nuclear Information System (INIS)

    Rodrigues, G.V.

    1986-01-01

    The theory of thermoluminescent emission, the characteristics of main thermoluminescent phosphorus, and some applications are presented. The main thermoluminescent dosemeters used in dosimetry are related. (M.C.K.)

  4. Thermoluminescent dosemeter in a X-ray diffractometer; Dosimetria termoluminiscente en un difractometro de rayos X

    Energy Technology Data Exchange (ETDEWEB)

    Mendoza A, D.; Gonzalez M, P.; Falcon B, T. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, Mexico D.F. (Mexico); Castano, V.M. [Instituto de Fisica, UNAM, Laboratorio Juriquilla, A.P. 1-1010, C.P. 76001, Queretaro (Mexico)

    1999-07-01

    In this work it was presented the results obtained of the dosimetry which was realized in a X-ray diffractometer for powders, trademark Siemens D5000, using the thermoluminescent signal generated by the X-rays in the commercial dosemeter TLD-100 of Harshaw, US. In according to the results obtained, the radiation quantity received by an analysed material in the diffractometer, will be proportional to exposure time and it can vary from unities until tenths of grays. These results are very outstanding when are analysed crystalline materials in a diffractometer, for knowing the present crystalline phases, mainly if these are highly sensitive to the ionizing radiation, as it is the case of the thermoluminescent materials. (Author)

  5. State of the art in thermoluminescent dosimetry using dosemeters with automated read-out

    International Nuclear Information System (INIS)

    Regulla, D.F.

    1979-01-01

    In the beginning, the results obtained with automated read-out TLD have been quite disappointing, both for manufacturers and users. In response to this, industry then seeked advice and assistance from those experienced in the practical use of dosemeters, and the development of the second generation of automated TLD has been a common task of manufacturers and purchosers. As a preliminary achievement of development work along this line, a dosemeter has been presented by a Japanese manufacturer, meeting the requirements in quite a satisfactory way. The device is characterized by a frequency of disturbance of -5 , a microprocessor for controlling the measuring and control functions, possibility of active input of device parameters and boundary conditions, extensive self-control of operating functions, complete data protection, and thin-film detector technique. Pursuing this line of development, and taking into account future recommentations, automated thermoluminescent dosimetry does seem to be a promising alternative to current measuring techniques in personnel monitoring. (orig./RW) [de

  6. Exposure levels to radiation in a nuclear medicine laboratory: measurements with thermoluminescence dosemeters

    International Nuclear Information System (INIS)

    Ruiz J, A.

    2000-01-01

    Since 1996 in the Nuclear medicine Department and another routine and research departments of the National Institute of Medical and Nutrition Sciences (INCMNSZ) are working at least with 12 radioactive sources opened and sealed. However, it was unknown if with the Tl personal dosemeters with lithium fluoride crystals (LiF), could be possible to receive information about the equivalent dose for each radionuclide or if there was some radionuclide that by being low energy emissor or beta energy emissor, it did not represent an ionization power sufficient to excite the crystals of the thermoluminescent dosemeters (DTL). In this work the obtained results with control dosemeters of diverse source opened or sealed are shown, which were put at 1 cm of the LiF crystals during a time exposure 18 hours. Therefore the objective of this work is to verify which sources excite or not to DTL crystals and taking the pertinent safety measures for each radionuclide. The conclusion is that the majority of dosemeters were excited by beta or gamma radiation of the radionuclides and that LiF crystals are able to receive equivalent doses until 1200 mSv, being this a guarantee for the staff, that in an any moment could be exposed to high dose in his working day. Also it was corroborated that the radionuclides more energizer are: iodine-131, iodine-125 and sodium-22 while the phosphorus-32 must be managed with careful, since the exposure to hands can result significant, just like the technetium-99m ( 99m Tc). (Author)

  7. Thermoluminescence and their applications

    International Nuclear Information System (INIS)

    Rodrigues, G.V.

    1986-01-01

    The theory involved in the phenomenon of thermoluminescence emission, the characteristics of the main thermoluminescent phosphorus and some of their applications are presented. Some of the main thermoluminescent dosemeters used in dosimetry are described. (M.C.K.) [pt

  8. Relative thermoluminescent efficiencies proton/gamma and helium/gamma of high temperature peaks in TLD-100 dosemeters

    International Nuclear Information System (INIS)

    Flores M, E.; Avila, O.; Rodriguez V, M.; Massillon, J.L.G.; Buenfil A, E.; Ruiz T, C.; Brandan, M.E.; Gamboa De Buen, I.

    2007-01-01

    This work presents measures of relative thermoluminescent efficiency of those high temperature peaks of TLD-100 dosemeters exposed to protons of 1.5 MeV and to helium nuclei of 3 and 7.5 MeV. A rigorous reading and of deconvolution protocol was used for the calculation of the TL efficiencies. Additionally an Excel program that facilitated the deconvolution adjustment process of the glow curves was elaborated. (Author)

  9. Electronic personal dosemeters

    International Nuclear Information System (INIS)

    Marshall, T.O.; Bartlett, D.T.; Burgess, P.H.; Cranston, C.S.; Higginbottom, D.J.; Sutton, K.W.

    1990-01-01

    Personal dosimetry services approved by their national authorities for category A workers, invariably use passive dosemeters incorporating photographic film or thermoluminescent detectors. However, the performance characteristics of electronic dosemeters has improved substantially over the past decade to such an extent that in the opening lecture of the Solid State Dosimetry Conference at Oxford in 1986 the development of an electronic 'smart card' based on a silicon detector was briefly discussed. This idea has been taken up and at least one development programme is in progress aimed at the production of an electronic dosemeter suitable for use as a legal device. The more important performance requirements of personal dosemeters for this purpose are discussed and the earlier electronic dosemeter designs and the latest devices under development to meet this specification are compared. (author)

  10. Design and validation of a single sphere multi-detector neutron spectrometer based on LiF: Mg,Cu,P thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Gomez-Ros, Jose Maria; Bedogni, Roberto; Moraleda, Montserrat; Romero, Ana; Delgado, Antonio; Esposito, Adolfo

    2010-01-01

    This communication describes a new neutron spectrometer consisting of pairs of 7 Li and 6 Li based thermoluminescent dosemeters (MCP-6, MCP-7) located at selected positions within a single moderating polyethylene sphere. The spatial arrangement of the dosemeters has been designed using the MCNPX Monte Carlo code to calculate the response matrix in order to obtain a nearly isotropic response for neutrons in the energy range up to 20 MeV. A partial validation of the calculated response matrix has been performed with the calibrated 241 Am-Be neutron source at the INFN-LNF Laboratory, using the shadow cone technique.

  11. Thermoluminescent dosemeters (TLD) exposed to high fluxes of gamma radiation, thermal neutrons and protons

    International Nuclear Information System (INIS)

    Gambarini, G.; Martini, M.; Meinardi, F.; Raffaglio, C.; Salvadori, P.; Scacco, A.; Sichirollo, A.E.

    1996-01-01

    Thermoluminescent dosemeters (TLD), widely experimented and utilized in personal dosimetry, have some advantageous characteristics which induce one to employ them also in radiotherapy. The new radiotherapy techniques are aimed at selectively depositing a high dose in cancerous tissues. This goal is reached by utilising both conventional and other more recently proposed radiation, such as thermal neutrons and heavy charged particles. In these inhomogeneous radiation fields a reliable mapping of the spatial distribution of absorbed dose is desirable, and the utilized dosemeters have to give such a possibility without notably perturbing the radiation field with the materials of the dosemeters themselves. TLDs, for their small dimension and their tissue equivalence for most radiation, give good support in the mapping of radiation fields. After exposure to the high fluxes of therapeutic beams, some commercial TL dosemeters have shown a loss of reliability. An investigation has therefore be performed, both on commercial and on laboratory made phosphors, in order to investigate their behaviour in such radiation fields. In particular the thermal neutron and gamma ray mixed field of the thermal column of a nuclear reactor, of interest for Boron Neutron Capture Therapy (B.N.C.T.) and a proton beam, of interest for proton therapy, were considered. Here some results obtained with new TL phosphors exposed in such radiation fields are presented, after a short description of some radiation damage effect on commercial LiF TLDs exposed in the (n th ,γ) field of the thermal column of a reactor. (author)

  12. Thermoluminescence albedo-neutron dosimetry

    International Nuclear Information System (INIS)

    Strand, T.; Storruste, A.

    1986-10-01

    The report discusses neutron detection with respect to dosimetry and compares different thermoluminescent dosimetry materials for neutron dosimetry. Construction and calibration of a thermoluminescence albedo neutron dosemeter, developed by the authors, is described

  13. Study of radioactive sources accumulation with application of thermoluminescence dosemeters on the base of alkaline earth metals sulfates

    International Nuclear Information System (INIS)

    Tokbergenov, I.; Sadykov, T.

    2001-01-01

    Methodic for study of accumulation and distribution of radioactive sources in a nature objects is developed. An essence of the method consists of in that quantity of accumulated radioactive sources in a nature objects is defining by absorption dose measured with help of thermoluminescent dosemeters on the base of alkaline earth metals sulfates such as CaSO 4 :Dy and SrSO 4 :Eu

  14. In-vivo dosimetry in radiotherapy: a comparison of the response of semiconductor and thermoluminescence (TLD700) dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Vynckier, S [Universite Catholique de Louvain, Brussels (Belgium). Cliniques Universitaires St. Luc; Greffe, J L; Loncol, T; Vanneste, F; Octave-Prignot, M; Denis, J M; De Patoul, N

    1995-12-01

    Semiconductor dosemeters and thermoluminescence dosemeters were calibrated in view of in-vivo dosimetry. Their response in a 8 MV photon beam and the respective correction factors for the treatment conditions were systematically studied. A total of 249 entrance and exit measurements with this dual detector combination were performed, mainly for treatments of the head and neck region. The resulting entrance and exit doses were compared with the expected doses at these positions, calculated on basis of the treatment and patient parameters. The results at the entrance showed a value of 1.010 (2.8% for the ratio of the measured to the calculated dose by diodes, 1.013) 4.9% for the ratio of the measured to the calculated dose by TLD and 1.003 (3.6% for the ratio of the measured dose by TLD to diodes. With respect the exit dose, the results were 0.998) 4.9%, 1.016 (7.7% and 1.019) 7.0% respectively after correction for the heterogeneity`s. Although the standard deviation for the TLD dosemeters is systematically larger than the standard deviation for the diodes, it is concluded that both dosemeters will yield similar results for-in-vivo dosimetry, if utilized under the same conditions.

  15. A contribution to thermoluminescence dosemeter system optimisation applied to environmental monitoring around nuclear power plants

    International Nuclear Information System (INIS)

    Dies, X.; Ortega, X.; Rosell, J.

    1990-01-01

    The factors that constitute a thermoluminescence dosimetric system have been studied and optimised, with the intention of defining strategies for improving precision and accuracy of TL dosimetry systems for environmental monitoring. An optimisation technique based on fractional factorial designs permitted a reduction in the number of necessary experiments and clarified interactions between various factors. The model for fading over time showed that TL response in LiF:Mg,Ti dosemeters was independent of time between irradiation and readout, and this was applied to fading corrections in nuclear power plant environmental monitoring. (author)

  16. About the measurements systems with pen and thermoluminescent dosemeters; Sobre los sistemas de medicion con dosimetros de pluma y los termoluminiscentes

    Energy Technology Data Exchange (ETDEWEB)

    Cortes I, M.E.; Ramirez G, F.P. [Instituto Mexicano del Petroleo, Eje Central Lazaro Cardenas 152. 07730 Mexico D.F. (Mexico)

    1998-12-31

    In this work it is presented dosimetric data obtained with pen and thermoluminescent dosemeters, which are used by the Occupational Exposure Personnel (OEP) of the Mexican Petroleum Institute (IMP)(1). It was marked several great characteristics as for example, the differences among units which use one and another dosemeter type. Likewise, it is given to know diverse problems that were had in the IMP at relating the data obtained with these dosemeters (which utilizes OEP) and the ICRP 60 recommendations 1990. One of the most important difficulties is to satisfy the recommended limits by ICRP, particularly those that are referring to the units and their complex calculations. With respect to the unities, the ICRP makes reference at the concepts `dose equivalent` and `effective dose` with the sievert unit, that the General Regulations for Radiological Safety associates with `dose equivalent` and `effective dose equivalent`. It was illustrated the type of dosimetric statistics which are obtained with the TLD lectures and a OEP pen dosemeter during 1997. (Author)

  17. Results of the EURATOM programme for comparison of individual dosemeters

    International Nuclear Information System (INIS)

    Julius, H.W.

    1976-01-01

    The results of the EURATOM dosemeter comparison-program for dosemeters used in the member states are given. Especially the results obtained in the Netherlands are examined and evaluated. The design and characteristics of the badge developed by the TNO-RD which are based on a thermoluminescent dosemeter are given

  18. Kerma determination in air on mamma by thermoluminescence

    International Nuclear Information System (INIS)

    Palacios P, L. L.; Rivera M, T.

    2009-10-01

    In this work the experimental results of the entrance exposition are shown and Kerma in air [mGy] in mamma obtained by irradiation of accreditation phantom of American College of Radiology (ACR). The irradiations were realized in a conventional mammography equipment of Hospital Juarez in Mexico; the technique used during the irradiations was of automatic exposition; the thickness for the phantom ACR obtained by the technique were of 4.2 and 4.5 cm; the kilo voltage pick was of 24 kV p , the time and the milli amperage per second variable. The measuring of Kerma in air was obtained with thermoluminescent dosemeters of solid state, of nano particles of zirconium dioxide prepared by the precipitation method. The dosemeters were homogenized previously in low energies of X-rays that are those used for mammography. The thermoluminescent dosemeters of ZrO 2 were calibrated by means of an ionization chamber for different expositions. The calibration curve is reported for the exposition and Kerma in air against thermoluminescent intensity obtained by reading of thermoluminescent dosemeters of ZrO 2 , as well as the technique employee for the Kerma determination in air and entrance exposition in mamma. (Author)

  19. Measurement of doses to aviator pilots using thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Azorin N, J.; Cruz C, D.; Rivera M, T.

    2004-01-01

    During the development of their work, the aviator pilots are exposed at high levels of natural radiation of bottom caused mainly by the cosmic radiation of galactic origin and lot. For such reason, the Metropolitan Autonomous University (UAM) and the Union Association of Aviator Pilots (ASPA), subscribed an agreement with the purpose of to measure the doses of ionizing radiation received by the aviator pilots of diverse air companies that man different types of airships and to determine if these doses surpass the one limit of 0.11 mSv/h settled down by the IAEA for the public in general; and if therefore, these workers should be considered as personnel occupationally exposed. In this work the obtained results when measuring the absorbed dose received by Mexican civil aviator pilots during the development of their work, using thermoluminescent dosemeters of LiF:Mg,Cu,P + Ptfe of national production are presented. The obtained results during the years of 2001 and 2002 show that the monthly doses received by the pilots surpass the one it limits established for the public in general, for what they should be considered as personnel occupationally exposed. (Author)

  20. Thermoluminescent Signals Caused by Disturbing Effects

    International Nuclear Information System (INIS)

    German, U.; Weinstein, M.; Ben-Shachar, B.

    1999-01-01

    One of the major sources of uncertainty in the measurement of low radiation doses by means of thermoluminescence dosemeters is the presence of disturbing thermoluminescence signals, especially luminescence caused by visible light, and by materials attached to the heated areas. Glow curves of thermoluminescence dosemeters contain useful information that can improve the accuracy and the reliability of the thermoluminescent measurements. The influence of the various disturbing effects can be recognised in the shape of the glow curves and can sometimes be separated from the exposure. Some examples are presented of signals arising from the two disturbing effects mentioned above, the signal contributed by Teflon used in the TLD-100 cards of Bicron/Harshaw and some abnormal glow curves due to dirt attached to the cards. Subtraction of the contributions due to these effects is suggested to obtain the net exposure signal. (author)

  1. Thermoluminescent response of dosemeters TLD-100 exposed to distinct baker protocols, irradiation and reading; Respuesta termoluminiscente de dosimetros TLD-100 sujetos a distintos protocolos de horneado, irradiacion y lectura

    Energy Technology Data Exchange (ETDEWEB)

    Estrada E, D

    2002-07-01

    The interest that motivated the realization of this work is the opposing discrepancy among the values of the efficiency thermoluminescent of heavy charged particles (PCP) regarding gamma radiation ({eta}{sub PCP}, {gamma}) reported in different works and gathered recently for Horowitz (Ho 01). The measures of relative thermoluminescent efficiency have importance in relation to the dosimetry associated to medical therapy with heavy charged particle. The measurement of {eta}{sub PCP}, {gamma} depends of multiple experimental factors. To understand this dependence it was quantified the effect of using different procedures experimental in the estimate of the relative thermoluminescent efficiency of dosemeters TLD-100 (LiF: Ti; Mg) for protons of 3 MeV regarding gamma rays of {sup 60}Co ({eta}{sub PCP}, {gamma}), varying the lot, the presentation, the one baked and the team reader. When was used the same global procedure values they were obtained for ({eta}{sub PCP}, {gamma}) inside the interval [0.30, 0.58] for the peak 5 and [0.35, 0.76] for the total thermoluminescent signal (Tl). At the use different equipment, baked and lot are obtained maximum differences of 30%, 12% and 6% respectively in the values of ({eta}{sub PCP}, {gamma}) for the peak 5, and 25%, 28% and 7% for the total Tl signal. The changes more significant of 36% and 44% for peak 5 and total signal respectively are obtained when changing the presentation of the dosemeter. As a complementary investigation it was studied the distribution of the Tl signal of each peak in the deconvolution of the curve of brightness with regard to the total signal in exposed dosemeters to high fluence of protons (1x10{sup 10} p/cm{sup 2}) varying the experimental protocol. It was found that the distribution of the Tl signal has a difference of the order of 5% when changing the used reading team. It was observed that the contribution of the peak 5 regarding the total signal are preserved when changing baked and that it

  2. Study of a new glass matrix by the thermoluminescence technique

    International Nuclear Information System (INIS)

    Ferreira, Pamela Z.; Vedovato, Uly P.; Cunha, Diego M. da; Dantas, Noelio O.; Silva, Anielle C.A.; Neves, Lucio P.; Perini, Ana P.; Carrera, Betzabel N.S.; Watanabe, Shigueo

    2015-01-01

    The thermoluminescence technique is widely used for both personal and for high-dose dosimetry. In this work, the thermoluminescence technique was utilized to study a new glass matrix, with nominal composition of 20Li 2 CO 3 .10Al 2 O 3 .20BaO.50B 2 O 3 (mol%), irradiated with different doses in a 60 Co source. The glow curves and the dose-response curve were obtained for radiation doses between 50 Gy and 900 Gy. The results showed that this new glass matrix presents potential use in high-dose dosimetry. (author)

  3. Obtention of thermoluminescent efficiencies by means of irradiation of TLD-100 dosemeters with proton beams helium and carbon

    International Nuclear Information System (INIS)

    Avila, O.; Rodriguez V, M.; Aviles, P.; Gamboa de Buen, I.; Buenfil, A.E.; Ruiz T, C.; Brandan, M.E.

    2002-01-01

    In this work, the advances of a serial of measurements of relative efficiency thermoluminescent of heavy charged particles (PCP) with respect to gamma radiation for TLD-100, dosemeters of LiF: Mg,Ti manufactured by the Harshaw-Bicron company are reported. The PCP are essentials in the implementation of dosimetry associated with medical applications. The measurements before gamma radiation were carrying out using the Vickrad irradiator of the National Institute of Nuclear Research at dose of 1.663 Gy. The measures which are reported about protons, helium and carbon were realized using the Pelletron accelerator of the Physics Institute of the UNAM. (Author)

  4. Clay as Thermoluminescence Dosemeter in diagnostic Radiology ...

    African Journals Online (AJOL)

    This paper reports the investigation of the basic thermoluminescence properties of clay at x-rays in the diagnostic radiology range, including dose monitoring in abdominal radiography. Clay sourced from Calabar, Nigeria, was tested for thermoluminescence response after irradiation at diagnostic radiology doses, including ...

  5. Optical readout method for solid-state dosemeters

    International Nuclear Information System (INIS)

    McDonald, J.C.; Eichner, F.N.; Stahl, K.A.; Miller, S.D.

    1986-07-01

    The readout of solid-state dosemeters is usually accomplished by heating to produce thermoluminescence. This technique has several disadvantages including stressing the dosemeter crystals, melting Teflon enclosures, and destroying the thin dosemeters designed for beta particle measurements. An optical readout method is being developed to avoid these difficulties. Standard dosemeters were irradiated to a dose of approximately 0.02 Sv with 137 Cs gamma rays. The dosemeters were then irradiated with light produced by a high-intensity xenon lamp. Various wavelength bands, from the ultraviolet through the visible and to the near-infrared, were used. The degree of trap emptying was found to be proportional to the total optical power incident. With the intensities used in the preliminary experiments, over 90% trap emptying was achieved. This new technique will prove useful for dosemeters that are encased in plastic for automatic processing. The details of this optical readout method, along with some possible applications in neutron and beta dosimetry are described. 7 refs., 3 figs

  6. Comparison of two models for the X-ray dispersion produced in a Novillo Tokamak with measurements make with thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Flores O, A.; Castillo, A.; Barocio, S.R.; Melendez L, L.; Chavez A, E.; Cruz C, G.J.; Lopez, R.; Olayo, M.G.; Gonzalez M, P.; Azorin N, J.

    1999-01-01

    It was presented the results to study about the X-ray dispersion produced in the Novillo Tokamak using thermoluminescent dosemeters (DTL). The measurements were make in the equatorial plane of Tokamak, along twelve radial directions. The dispersion is observed due to the radiation interaction with walls surrounding the machine. It was proposed two types of heuristic mathematical methods for describing the X-ray dispersion, comparing them with the experimental data obtained with Dtl. The predictions of both models are adjusted well to the experimental data. (Author)

  7. Electron scattering effects on absorbed dose measurements with LiF-dosemeters

    International Nuclear Information System (INIS)

    Bertilsson, G.

    1975-10-01

    The investigation deals with absorbed dose measurements with solid wall-less dosemeters. Electron scattering complicates both measurement of absorbed dose and its theoretical interpretation. The introduction of the dosemeter in a medium causes perturbations of the radiation field. This perturbation and its effect on the distribution of the absorbed dose inside the dosemeter is studied. Plane-parallel LiF-teflon dosemeters (0.005 - 0.1 g.cm -2 ) are irradiated by a photon beam ( 137 Cs) in different media. The investigation shows that corrections must be made for perturbations caused by electron scattering phenomena. Correction factors are given for use in accurate absorbed dose determinations with thermoluminescent dosemeters. (Auth.)

  8. Determination of the dose index in computerized tomography using thermoluminescent dosemeters; Determinacion del indice de dosis en tomografia computada usando dosimetros termoluminiscentes

    Energy Technology Data Exchange (ETDEWEB)

    Azorin, J.C.; Calderon, A. [Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada-IPN, 11500 Mexico D.F. (Mexico); Azorin, J. [UAM-I, 09340 Mexico D.F. (Mexico)

    2004-07-01

    In this work the obtained results of the determination of the dose index are presented in thorax studies in computed tomography and helical tomography carried out in Mexico using thermoluminescent dosemeters of LiF: Mg,Cu,P + Ptfe developed and manufactured in our country. The results showed that under similar conditions of irradiation and operation (pitch = 1), significant differences don't exist among the doses absorbed measures in the phantom due to the two types of used tomographs. (Author)

  9. Prediction analysis of dose equivalent responses of neutron dosemeters used at a MOX fuel facility

    International Nuclear Information System (INIS)

    Tsujimura, N.; Yoshida, T.; Takada, C.

    2011-01-01

    To predict how accurately neutron dosemeters can measure the neutron dose equivalent (rate) in MOX fuel fabrication facility work environments, the dose equivalent responses of neutron dosemeters were calculated by the spectral folding method. The dosemeters selected included two types of personal dosemeter, namely a thermoluminescent albedo neutron dosemeter and an electronic neutron dosemeter, three moderator-based neutron survey meters, and one special instrument called an H p (10) monitor. The calculations revealed the energy dependences of the responses expected within the entire range of neutron spectral variations observed in neutron fields at workplaces. (authors)

  10. Application of beta transmission for quality assurance of TLD tape dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Nagpal, J.S.; Gupta, S.K.; Supe, S.J. (Bhabha Atomic Research Centre, Bombay (India). Div. of Radiological Protection)

    1993-03-01

    Transmission of [sup 204]Tl [beta] radiation enables measurement of differences of [+-] 2 mg/cm[sup 2] in thermoluminescent tape dosemeter thickness. This thickness variation gives rise to a maximum variation of [+-] 5% in the TL output per unit area of the dosemeters. Measurements of [beta] transmission on dosemeter samples drawn from several batches of skived tape as well as those prepared by compressing technique have given consistent results. The [beta] transmission method itself can be used for quality assurance and selection of a uniform set of dosemeters over thickness range 30-150 mg/cm[sup 2]. (author).

  11. Thermoluminescent dosemeters of CaSO4: Dy + Teflon

    International Nuclear Information System (INIS)

    Campos, L.L.

    1989-01-01

    The development of a pellet dosemeter of CaSO 4 : Dy + Teflon at the Dosimetric Materials Production Laboratory - IPEN/Brazilian CNEN-SP is presented. The pellets were produced by cold pressing and sintering a mixture of CaSO 4 : Dy and Teflon powders. The pellet characteristics from the point of view of dosimetry. A filter combination providing an energy independent response from 20 KeV to 1,25 MeV was obtained. The dosemeter consists of three pellets sealed between two thin plastic sheets and placed under plastic and lead filters. The combination of these three filters allows to determine the energy of an unknown source. (author) [pt

  12. The fast neutron response of 7LiF thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Knipe, A.D.

    1990-02-01

    A series of experiments was performed in 1982 to determine the neutron energy response of the 7 LiF thermoluminescent dosemeter (TLD) employed in European zero power fast reactor gamma-ray energy deposition studies. Preliminary results of this work were included in a 1985 International experimental data-base of TLD neutron sensitivities and provided the most consistent set of data within the 7 LiF compilation. A more detailed interpretation of these data has improved the results still further, giving a mean relative neutron to gamma efficiency of 0.118±0.005. The main objective of this re-evaluation was to establish recommended neutron energy response values to replace a 1974 data set for the analysis of in-core fast reactor measurements. This was achieved by combining the mean experimental relative efficiency of 0.118 with calculated energy dependent kerma factors. The kerma factors for the TLD were based on US National Bureau of Standards values and a composition determined by chemical analysis. Adoption of the revised neutron energy response data set produces a small increase in the measured gamma-ray energy deposition of typically 2% relative to the 1974 data. However, more importantly, the detailed analysis of the experimental response data has significantly improved confidence in the neutron corrections applied to in-core TLD gamma-ray energy deposition measurements. (author)

  13. Obtention of a thermoluminescent material for dosimetry of ionizing radiation

    International Nuclear Information System (INIS)

    Gonzalez M, P.R.

    1990-01-01

    The thermoluminescent dosemeters are small crystals which suffer changes in their structure by the radiation effect, being displaced the electrons toward higher energy levels. On heating the previously irradiated crystals, the electrons come back to their base state emitting light photons. The light quantity emitted is proportional to the received radiation dose. The light quantity emitted is proportional to the received radiation dose. The lithium fluoride is one of the thermoluminescent materials considered as tissue equivalents by having a low effective atomic number (Z ef ). At present, the more used commercial product used of this type is the TLD-100*. In this work the obtained results in the preparation of the lithium fluoride thermoluminescent material are presented. This is activated with magnesium (Mg) and titanium (Ti), which we have labelled as: LiF: Mg, Ti. The results from the tests performed for verifying his thermoluminescent properties are presented too, as powder form as in pellets form. These tests were performed in simultaneous form with TLD-100 samples, which is considered as reference. The LiF: Mg, Ti thermoluminescent material manufactured in the ININ presents similar dosimetric characteristics to those ones of the TLD-100. Therefore being able to replace the imported dosemeters. * (TLD-100 is a commercial trademark registered by Harshaw/Filtrol (US) for LiF: Mg, Ti Tl dosemeters) (Author)

  14. Environmental gamma-ray dose measurements with thermoluminescence dosemeters (TLD) and environmental radiation characteristics

    International Nuclear Information System (INIS)

    Kanematsu, Seiko

    1999-01-01

    It is important to evaluate environmental gamma-ray exposure both at work and home in order to assess people's collective dosages. Environmental gamma radiation was measured for air-absorbed dose with a thermoluminescence dosemeter at various points in the workplace and Ningyotoge, and workplace radiation characteristics were analyzed. From the results, the public dose due to gamma rays generated artificially was assessed to be sufficiently lower than the annual limit. For indoor environments of the workplace, the maximum dosage rate among measured values was 97 nGy/h and the minimum value was 70 nGy/h, the average over one year was 83 nGy/h. The average annual outdoor dosage for a year was 82 nGy/ h. In Ningyotoge, the maximum was 103 nGy/h, minimum 60 nGy/h, and average 88 nGy/h. These values depend on the nature of the soil and weather factors, showing higher values in the summer than in the winter in the workplace. There was no significant difference in the dosage rate in houses and the workplace. (author)

  15. Thermoluminescent and optical processes in alkaline halogenides dosemeters contaminated with Europium; Procesos opticos y termoluminiscentes en dosimetros de halogenuros alcalinos contaminados con Europio

    Energy Technology Data Exchange (ETDEWEB)

    Barboza F, M.; Melendrez, R.; Castaneda, B.; Pedroza M, M.; Chernov, V.; Perez S, R.; Aceves, R. [Centro de Investigacion en Fisica, Universidad de Sonora, A.P. 5-088, 83190 Hermosillo, Sonora (Mexico)

    2000-07-01

    Recent research results are presented about the properties of the optical processes of photo transferred thermoluminescence (TLFT), optical whitening (BO), thermoluminescence induced by light (TLL) and its effect in the thermoluminescent curve (Tl) produced by ionizing and non-ionizing radiation. The systematic analysis of all these processes, acquires a singular importance due that actually the alkaline halogenide crystals are object of intense investigations which analyse their potential applications as detectors and radiation dosemeters through stimulated optical luminescence techniques or thermoluminescence. The obtained data show that the Tl curve of material with this nature can be enormously affected by exposure of phosphorus to the environmental light or UV. This is in part due to liberation processes of charge bearers are shouted and makes a subsequent trapping in less temperature traps; at the same time that induce changes in the intensity of determined Tl bands. Additionally, also it is observed that mentioned phenomena are related as with wavelength of incident light as of the illumination time. Finally, the obtained information allows to conclude that although the illumination effect is extremely complex, it is associated and can be explained mainly with phenomena that implicate the electrons excitation trapped in form of F centers and trapping mechanisms or radioactive and non-radioactive recombination. (Author)

  16. A new, passive dosemeter for gamma, beta and neutron radiations

    Energy Technology Data Exchange (ETDEWEB)

    Jones, L A; Stokes, R P, E-mail: rpstokes@dstl.gov.uk [Defence Science and Technology Laboratory, Environmental Sciences Department, Alverstoke, Gosport, Hants, PO12 2DL (United Kingdom)

    2011-03-01

    The Defence Science and Technology Laboratory (Dstl) provides personal radiation dosimetry to the UK Ministry of Defence. Dstl has recently developed a dosemeter that is based on a combination of thermoluminescent and etched-track detectors. The Dstl Combined Dosemeter is capable of assessing doses due to photons, beta particles and neutrons. This paper presents the laboratory type testing results for the Combined Dosemeter, and also describes the procedure for calibrating the dosemeter for use in workplace neutron fields. The Combined Dosemeter meets the type test requirements that are relevant to its intended applications, and gives neutron doses that are within 50% of the true dose in the workplaces in which it is used, even when the wearer has the potential to be exposed to a variety of neutron spectra (e.g. on board nuclear-powered submarines).

  17. A new, passive dosemeter for gamma, beta and neutron radiations

    International Nuclear Information System (INIS)

    Jones, L A; Stokes, R P

    2011-01-01

    The Defence Science and Technology Laboratory (Dstl) provides personal radiation dosimetry to the UK Ministry of Defence. Dstl has recently developed a dosemeter that is based on a combination of thermoluminescent and etched-track detectors. The Dstl Combined Dosemeter is capable of assessing doses due to photons, beta particles and neutrons. This paper presents the laboratory type testing results for the Combined Dosemeter, and also describes the procedure for calibrating the dosemeter for use in workplace neutron fields. The Combined Dosemeter meets the type test requirements that are relevant to its intended applications, and gives neutron doses that are within 50% of the true dose in the workplaces in which it is used, even when the wearer has the potential to be exposed to a variety of neutron spectra (e.g. on board nuclear-powered submarines).

  18. Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study.

    Science.gov (United States)

    Koivisto, J; Schulze, D; Wolff, J; Rottke, D

    2014-01-01

    The objective of this study was to compare the performance of metal oxide semiconductor field-effect transistor (MOSFET) technology dosemeters with thermoluminescent dosemeters (TLDs) (TLD 100; Thermo Fisher Scientific, Waltham, MA) in the maxillofacial area. Organ and effective dose measurements were performed using 40 TLD and 20 MOSFET dosemeters that were alternately placed in 20 different locations in 1 anthropomorphic RANDO(®) head phantom (the Phantom Laboratory, Salem, NY). The phantom was exposed to four different CBCT default maxillofacial protocols using small (4 × 5 cm) to full face (20 × 17 cm) fields of view (FOVs). The TLD effective doses ranged between 7.0 and 158.0 µSv and the MOSFET doses between 6.1 and 175.0 µSv. The MOSFET and TLD effective doses acquired using four different (FOV) protocols were as follows: face maxillofacial (FOV 20 × 17 cm) (MOSFET, 83.4 µSv; TLD, 87.6 µSv; -5%); teeth, upper jaw (FOV, 8.5 × 5.0 cm) (MOSFET, 6.1 µSv; TLD, 7.0 µSv; -14%); tooth, mandible and left molar (FOV, 4 × 5 cm) (MOSFET, 10.3 µSv; TLD, 12.3 µSv; -16%) and teeth, both jaws (FOV, 10 × 10 cm) (MOSFET, 175 µSv; TLD, 158 µSv; +11%). The largest variation in organ and effective dose was recorded in the small FOV protocols. Taking into account the uncertainties of both measurement methods and the results of the statistical analysis, the effective doses acquired using MOSFET dosemeters were found to be in good agreement with those obtained using TLD dosemeters. The MOSFET dosemeters constitute a feasible alternative for TLDs for the effective dose assessment of CBCT devices in the maxillofacial region.

  19. Investigation of dosimetric characteristics of the high sensitivity LiF:Mg,Cu,P Thermoluminescent Dosemeter and its applications in diagnostic radiology - a review

    International Nuclear Information System (INIS)

    Fung, K.L.

    2004-01-01

    This study investigated the dosimetric properties of the high sensitivity TLD (Thermoluminescent Dosemeter) of LiF:Mg,Cu,P and its applications in diagnostic radiology. A reproducible readout and annealing regime for this high sensitivity TLD was developed in the initial part of this study with the newly installed automatic TLD Reader system. Basic dosimetric characteristics of this T.L. dosemeter were then investigated. This paved the foundation for subsequent selected novel application studies in diagnostic radiology. This study exploits the favourable dosimetric properties of these T.L. dosemeters in some selected novel dosimetric applications in diagnostic radiology with an anthropomorphic phantom. The applications studied in radiological procedures included: dose reduction in lumbar spine radiography utilizing the 'anode heel effect'; gonad dose variation with kV p in chest radiography; foetal dose comparison between computed tomography (CT) and computed radiography (CR) in X-ray pelvimetry; lens dose reduction with bismuth eye-shields in CT brain studies; foetal dose assessment of early pregnancy in common high risk radiological examinations. It is anticipated that the unique and favourable dosimetric performance of LiF:Mg,Cu,P T.L. phosphor will be exploited further in measurements of low level dose received by patients and staff in diagnostic radiological procedures such as paediatric X-ray examinations

  20. Investigation of dosimetric characteristics of the high sensitivity LiF:Mg,Cu,P Thermoluminescent Dosemeter and its applications in diagnostic radiology - a review

    Energy Technology Data Exchange (ETDEWEB)

    Fung, K.L. E-mail: orkarl@polyu.edu.hk

    2004-05-01

    This study investigated the dosimetric properties of the high sensitivity TLD (Thermoluminescent Dosemeter) of LiF:Mg,Cu,P and its applications in diagnostic radiology. A reproducible readout and annealing regime for this high sensitivity TLD was developed in the initial part of this study with the newly installed automatic TLD Reader system. Basic dosimetric characteristics of this T.L. dosemeter were then investigated. This paved the foundation for subsequent selected novel application studies in diagnostic radiology. This study exploits the favourable dosimetric properties of these T.L. dosemeters in some selected novel dosimetric applications in diagnostic radiology with an anthropomorphic phantom. The applications studied in radiological procedures included: dose reduction in lumbar spine radiography utilizing the 'anode heel effect'; gonad dose variation with kV{sub p} in chest radiography; foetal dose comparison between computed tomography (CT) and computed radiography (CR) in X-ray pelvimetry; lens dose reduction with bismuth eye-shields in CT brain studies; foetal dose assessment of early pregnancy in common high risk radiological examinations. It is anticipated that the unique and favourable dosimetric performance of LiF:Mg,Cu,P T.L. phosphor will be exploited further in measurements of low level dose received by patients and staff in diagnostic radiological procedures such as paediatric X-ray examinations.

  1. Fiber-optic-coupled dosemeter for remote optical sensing of radiation

    International Nuclear Information System (INIS)

    Justus, B.L.; Huston, A.L.

    1996-01-01

    Remote sensing technologies for the detection and measurement of ionizing radiation exposure are of current interest for applications such as patient dose verification during radiotherapy and the monitoring of environmental contaminants. Fiberoptic-based sensing is attractive due to the advantages of small size, low cost, long life and freedom from electromagnetic interference. Several fiberoptic-based radiation sensing systems have been described that utilize radiation induced changes in the optical characteristics of the fiber such as reduced transmission as a result of darkening of the glass, optical phase shifts due to heating, or changes in the birefringence of a polarization-maintaining fiber. The measurement of radiation induced darkening is limited in both sensitivity and dynamic range and requires long fiber lengths. Phase shift measurements require the use of single-mode lasers, phase sensitive interferometric detection, long fiber lengths and complex signal processing techniques. Alternatively, thermoluminescent (TL) phosphor powders have been coated onto fiberoptic cables and remote dosimetry measurements performed using traditional laser heating techniques. The sensitivity is limited by the requirement for a very thin layer of phosphor material, due to problems associated with light scattering and efficient heating by thermal diffusion. In this paper we report the development of an all-optical, fiber-optic-coupled, thermoluminescence dosemeter for remote radiation sensing that offers significant advantages compared to previous technologies. We recently reported the development of an optically transparent, TL glass material having exceptionally good characteristics for traditional dosimetry applications. We also reported a modified TL glass incorporating a rare earth ion dopant in order to absorb light from a semiconductor laser and utilize the absorbed light energy to internally heat the glass and release the trapped electrons. (author)

  2. Characteristics of dosemeter types for skin dose measurements in practice

    International Nuclear Information System (INIS)

    Van, D. J.; Bosmans, H.; Marchal, G.; Wambersie, A.

    2005-01-01

    A growing number of papers report deterministic effects in the skin of patients who have undergone interventional radiological procedures. Dose measurements, and especially skin dose measurements, are therefore increasingly important. Methods and acceptable dosemeters are, however, not clearly defined. This paper is the result of a literature overview with regard to assessing the entrance skin dose during radiological examinations by putting a dosemeter on the patient's skin. The relevant intrinsic characteristics, as well as some examples of clinical use of the different detector types, are presented. In this respect, thermoluminescence, scintillation, semiconductor and film dosemeters are discussed and compared with respect to their practical use. (authors)

  3. Evaluation of glasses containing cadmium for high dose dosimetry by the thermoluminescence technique

    International Nuclear Information System (INIS)

    Carvalho, Gabriel Soares Marchiori de; Ferreira, Pamela Zati; Cunha, Diego Merigue da; Dantas, Noelio Oliveira; Silva, Anielle C.A.; Perini, Ana Paula; Neves, Lucio Pereira; Carrera, Betzabel Noemi Silva; Watanabe, Shigueo

    2016-01-01

    New glass matrices were evaluated for high dose dosimetry by the thermoluminescence technique. Their nominal composition are 20Li_2CO_3.10Al_2O_3.15CdO.55B_2O_3 and 20Li_2CO_3.10Al_2O_3.20CdO.50B_2O_3 (mol%). The glass matrices were irradiated with different doses: 50, 100, 200, 500, 700 and 900 Gy, and the thermoluminescence emission curves were obtained for each of these values. The results show a great potential of using these matrices in high dose dosimetry. (author)

  4. Infrared luminescence and thermoluminescence of lithium borate glasses doped with Sm3+ ions

    Directory of Open Access Journals (Sweden)

    Anjaiah J.

    2015-03-01

    Full Text Available Thermoluminescence (TL characteristics of X-ray irradiated pure and doped with Sm3+ ions Li2O-MO-B2O3 (where MO=ZnO, CaO, CdO glasses have been studied in the temperature range of 303 to 573 K. All the pure glasses exhibited single TL peaks at 382 K, 424 K and 466 K. When these glasses were doped with Sm3+ ions no additional peaks have been observed but the glow peak temperature of the existing glow peak shifted gradually towards higher temperatures with gain in intensity of TL light output. The area under the glow curve was found to be maximum for Sm3+ doped glasses mixed with cadmium oxide as a modifier. The trap depth parameters associated with the observed TL peaks have been evaluated using Chen’s formulae. The possible use of these glasses in radiation dosimetry has been described. The results clearly showed that samarium doped cadmium borate glass has a potential to be considered as a thermoluminescence dosimeter.

  5. Relative thermoluminescent efficiencies proton/gamma and helium/gamma of peaks of high temperature in TLD-100 dosemeters

    International Nuclear Information System (INIS)

    Flores M, E.

    2007-01-01

    The increase of the applications of ion beams in radiotherapy treatments has generated interest in the study of the thermoluminescent materials (TL) that allow to determine the applied doses. A way to quantify the TL response from these materials to ions is by means of the relative thermoluminescent efficiency. In the group of Thermoluminescent dosimetry of the Institute of Physics of the UNAM (IFUNAM) the thermoluminescent response of the TLD-100 dosemeters has been studied, which present a glow curve characteristic with several peaks that correspond to traps and luminescent centers in the material. The stable peaks know each other as 4, 5, 6a, 6b, 7, 8, 9 and 10. The efficiencies should be measured using the response so much to the radiation of interest (in this case protons and helium ions) as the response to gamma radiation. In previous works with ions of low energy taken place in the Pelletron accelerator of the IFUNAM was only measured the TL efficiency for the peak 5 and the total signal. It had not been possible to measure the efficiency of the peaks of high temperature (6a-10) because, for the gamma radiation, the peaks of high temperature show very small signals; however, recently Massillon carries out measures of efficiency TL of peaks of high temperature for ions of intermediate energy using a protocol special of reading and of deconvolution that allows to measure the signals coming from the peaks of high temperature. In this work is implemented this same protocol to complete the study of TL efficiencies at low energy of protons and helium and to determine if the values of efficiency depend on the used reading protocol. For it is reported it measures of the relative efficiency of the peaks of high temperature from the TLD-100 exposed to protons of 1.5 MeV and nuclei of helium of 3 and 7.5 MeV. (Author)

  6. Thermoluminescent dosemeters characterization for patient dosimetry in diagnostic radiology preliminary results;Caracterizacao de dosimetros termoluminescentes para dosimetria de pacientes em radiodiagnostico - resultados iniciais

    Energy Technology Data Exchange (ETDEWEB)

    Castro, William J.; Squair, Peterson L. [Faculdade Novo Rumo, Belo Horizonte, MG (Brazil); Centro de Desenvolvimento de Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil); Gonzaga, Natalia B. [Universidade Unincor, Belo Horizonte, MG (Brazil); Centro de Desenvolvimento de Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil); Nogueira, Maria S.; Silva, Teogenes A. da [Centro de Desenvolvimento de Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2009-07-01

    The determination of the metrological characteristics of thermoluminescent (TL) dosimeters plays an important role in dosimetry of patients submitted to x-ray examinations for diagnostic purpose. Entrance surface doses can be measured with TL dosimeters to verify the compliance with the diagnostic reference levels. Organ doses can be estimated through TL measurements in an anthropomorphic phantom which it allows the radiation risk assessment. In this work, LiF:Mg,Ti (TLD-100) rod and chip thermoluminescent dosimeters were characterized considering their use for patient dosimetry in computerized tomography and mammography. Preliminary results showed that TL dosimeters have a response reproducibility of 7.8% and 4.8% and homogeneity of 18.4% and 6.5% for rod and chip shapes, respectively. (author)

  7. Thermoluminescent response of LiF before variation of the heating rate

    International Nuclear Information System (INIS)

    Barrios, R.; Avila, O.

    2003-01-01

    Comparisons of glow curves of lithium fluoride dosemeters TLD-100 measured to two heating rates with the purpose of quantifying the change in the temperature of the peaks 5 and 7 for the thermoluminescent reader equipment Harshaw 4000 of the thermoluminescence laboratory of the ININ were carried out. (Author)

  8. Dose equivalent response of personal neutron dosemeters as a function of angle

    International Nuclear Information System (INIS)

    Tanner, J.E.; McDonald, J.C.; Stewart, R.D.; Wernli, C.

    1997-01-01

    The measured and calculated dose equivalent response as a function of angle has been examined for an albedo-type thermoluminescence dosemeter (TLD) that was exposed to unmoderated and D 2 O-moderated 252 Cf neutron sources while mounted on a 40 x 40 15 cm 3 polymethylmethacrylate phantom. The dosemeter used in this study is similar to many neutron personal dosemeters currently in use. The detailed construction of the dosemeter was modelled, and the dose equivalent response was calculated, using the MCNP code. Good agreement was found between the measured and calculated values of the relative dose equivalent angular response for the TLD albedo dosemeter. The relative dose equivalent angular response was also compared with the values of directional and personal dose equivalent as a function of angle published by Siebert and Schuhmacher. (author)

  9. Evaluation of glasses containing cadmium for high dose dosimetry by the thermoluminescence technique

    Energy Technology Data Exchange (ETDEWEB)

    Carvalho, Gabriel Soares Marchiori de; Ferreira, Pamela Zati; Cunha, Diego Merigue da; Dantas, Noelio Oliveira; Silva, Anielle C.A.; Perini, Ana Paula; Neves, Lucio Pereira, E-mail: lucio.neves@ufu.br [Universidade Federal de Uberlandia (INFIS/UFU), MG (Brazil). Instituto de Fisica; Caldas, Linda V.E. [Instituto de Pesquisas Energeticas e Nucleares (IPEN-CNEN/SP), Sao Paulo, SP (Brazil); Carrera, Betzabel Noemi Silva; Watanabe, Shigueo [Universidade de Sao Paulo (IF/USP), SP (Brazil). Instituto de Fisica

    2016-07-01

    New glass matrices were evaluated for high dose dosimetry by the thermoluminescence technique. Their nominal composition are 20Li{sub 2}CO{sub 3}.10Al{sub 2}O{sub 3}.15CdO.55B{sub 2}O{sub 3} and 20Li{sub 2}CO{sub 3}.10Al{sub 2}O{sub 3}.20CdO.50B{sub 2}O{sub 3} (mol%). The glass matrices were irradiated with different doses: 50, 100, 200, 500, 700 and 900 Gy, and the thermoluminescence emission curves were obtained for each of these values. The results show a great potential of using these matrices in high dose dosimetry. (author)

  10. Study of a new glass matrix by thermoluminescent technique for high-dose dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Ferreira, Pamela Z.; Carvalho, Gabriel S. Marchiori de; Cunha, Diego M. da; Dantas, Noelio O.; Silva, Anielle C.A.; Neves, Lucio P.; Perini, Ana P., E-mail: anapaula.perini@ufu.br [Universidade Federal de Uberlandia (UFU), MG (Brazil). Instituto de Fisica; Linda, V.E. Caldas [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Carrera, Betzabel N.S.; Watanabe, Shigueo [Universidade de Sao Paulo (IF/USP), Sao Paulo, SP (Brazil). Instituto de Fisica

    2016-07-01

    The thermoluminescence technique is widely used for both personal and for high-dose dosimetry. In this work, the thermoluminescence technique was utilized to study a new glass matrix, with nominal composition of 20Li{sub 2}CO{sub 3}.10Al{sub 2}O{sub 3}.30BaO.40B{sub 2}O{sub 3} (mol%), irradiated with different doses in a {sup 60}Co source. The glow curves and the dose-response curve were obtained for radiation doses of 10, 50, 100, 200 e 700 Gy. The results showed that this new glass matrix has potential use in high-dose dosimetry. (author)

  11. An optical fibre-type silicate glass thermoluminescent detector

    International Nuclear Information System (INIS)

    Zheng Zheng; Dai Honggui; Hu Shangze; Liu Jian; Fang Jie

    1991-01-01

    A description of dosimetric properties and the preparation method of an optical fibre-type silicate glass thermoluminescent detector (TLD) is presented. Results showed that this new phosphor is a good one which could be used as a routine dosimeter in the range 10 -1 -10 3 Gy. The preparation method is a new one which differs greatly from all previous ones. Furthermore this kind of detector is small and of low weight. (orig.)

  12. The use of active personal dosemeters as a personal monitoring device: Comparison with TL dosimetry

    International Nuclear Information System (INIS)

    Boziari, A.; Koukorava, C.; Carinou, E.; Hourdakis, C. J.; Kamenopoulou, V.

    2011-01-01

    The use of active personal dosemeters (APDs) not only as a warning device but also, in some cases, as an official and hence stand-alone dosemeter is rapidly increasing. A comparison in terms of dose, energy and angle dependence, among different types of APD and a routinely used whole-body thermoluminescence dosemeter (TLD) has been performed. Significant differences were found between the TLD readings and mainly some not commonly used APDs. The importance of choosing the best adapted APD according to the radiation field characteristics is pointed out. (authors)

  13. An intercomparison of the thermoluminescent efficiency of various preparations of lithium fluoride

    CERN Document Server

    Driscoll, C M H

    1977-01-01

    Possible alternative sources of supply of lithium fluoride to that used at present in the NRPB automated TLD system dosemeter are being considered on the grounds of economy and quality control. Initial measurements on the thermoluminescent properties and efficiency of various preparations of lithium fluoride are described, including the Harshaw TLD-700 which is currently in use in the automated thermoluminescent dosemeter. Material obtained from British Nuclear Fuels Ltd (which may be an atypical batch) showed 75% of the sensitivity of the Harshaw material, with around 20% fading of the signal after 1 month's storage. French material, obtained from Desmarquest and Carbonisation Entreprise et Ceramique and used by the Commissariat a l'Energie Atomique, showed good fading resistance but, after standard preparational treatments, was only half as sensitive in the thermoluminescent response. Without additional treatments to improve the characteristics of these materials, they cannot be considered as being satisfac...

  14. Intercomparison of Environmental Dosemeters Using Various TL Materials and Dosimetry Systems

    International Nuclear Information System (INIS)

    Crnic, B.; Gobec, S.; Zorko, B.; Knezevic, Z.; Majer, M.; Ranogajec-Komor, M.

    2013-01-01

    The aim of the present work was to compare the ambient dose equivalent (H*(10)) values determined at 20 sites around NPP Krško, using different thermoluminescence (TL) materials and various dosimetry systems. The H*(10) was measured by the CaF 2 :Mn (TLD-400) provided by the Jozef Stefan Institute (JSI) Ljubljana, Slovenia. These dosemeters were deployed in the environment in plastic bags and suspended inside the plastic bottles. On the other hand the Ruder Boskovic Institute (RBI), Zagreb, Croatia applied LiF:Mg,Cu,P (TLD-100H), CaF 2 :Mn, Al 2 O 3 :C TL detectors and radiophotoluminescence (RPL) glass dosemeters type SG1. They were placed at the same locations in as much as possible same conditions as JSI detectors. According to the protocol established for this intercomparison, the control and transport detectors (not deployed in the environment) were held in dark storage containers and used to determine the background radiation. The TL responses were corrected for individual sensitivity of the TL detectors which is an important factor after the calibration irradiations. The calibration irradiations were performed by 137Cs sources provided in Secondary Standard Dosimetry Laboratory facilities at the JSI and RBI. The results obtained by different TL materials and different dosimetry systems show interesting features especially concerning local environmental peculiarities. The H*(10) obtained by the dosemeters of various types deployed in the countryside fluctuate less than 10 %. The outcome should emphasize also in the manner that the results obtained and reported in the intercomparison are traceable to the primary standards.(author)

  15. The new IRSN passive dosemeter using the RPL technique

    International Nuclear Information System (INIS)

    Fraboulet, P.; Cale, E.; Itie, C.; Bottollier-Depois, Jean-Francois

    2008-01-01

    Full text: The Institute for Radiological Protection and Nuclear Safety (IRSN) is the French public organisation in charge of research and expertise into nuclear safety and radiation protection. IRSN provides also services in these areas like dosimetric survey for 150,000 workers liable to be exposed to ionising radiation. At present whole-body dosemeters provided by the IRSN laboratory for photons and beta-particles are based on photographic films. In order to anticipate likely supplying decrease and to improve the service, IRSN has decided to replace photographic films by dosemeters based on the radio-photoluminescence (RPL) technology, supplied by the Japanese Company Chiyoda Technol. RPL was chosen after a thorough comparison with other available passive techniques (thermoluminescent dosemeters: TLD, and optically stimulated luminescence: OSL). Constraints due to large-scale production of dosemeters (packaging, shipping, etc.) were taken into account as well. Irradiation tests for five different dosemeters, based on the 3 available techniques (TLD, OSL, RPL) were performed. The main results of these tests are presented (energy, dose and angular response, detection threshold, etc.) together with main technical characteristics of the new RPL dosemeter (range in energy and dose, homogeneity, reproducibility, re-reading, information about radiation type from reading, etc.). This dosemeter is expected to give significative improvements in the IRSN laboratory's provision of service thanks to better dosimetric performances and because it offers the possibility to get information to analyse over-exposure circumstances. (author)

  16. Thermoluminescence properties of alkali borate glasses containing neodymium

    Energy Technology Data Exchange (ETDEWEB)

    Abbas, A.F.; Henaish, B.A.; Kenaway, M.A.; Salem, L.R.

    1988-01-01

    The thermoluminescence properties of sodium borate glasses as a function of neodymium oxide content as well as the divalent metal oxides (RO = ZnO, MgO and CaO) in replacement of Na/sub 2/O have been investigated. It is observed that the addition of Nd/sub 2/O/sub 3/ imparts to the host glass a monopeak glow curve according to an active luminescent centre (E approx. = 0.97 eV to 1.232 eV). The gradual addition of neodymium oxide to the sodium borate glass causes gradual enhancement in the TL-intensity up to a quenching concentration value (4 g Nd/sub 2/O/sub 3/ added to 100 g of borate glass) above which a draw back in TL-intensity occurs. On the other hand the replacement of 5 wt% Na/sub 2/O by RO shows that CaO dominates the other two divalent metal oxides used, as it possesses a much deeper luminescent trap (1.232 eV). The results obtained suggest that these glasses can be used in radiation detection and dosimetry. The ..gamma..-induced Tl-signal of such type of glass is found to be reproducible within an acceptable error of not more than 3.5% in all individual and group scattering over the detector samples each of which is used 10 times for evaluating the same ..gamma..-dose.

  17. Mechanical properties of a co-extruded Metallic Glass/Alloy (MeGA) rod-Effect of the metallic glass volume fraction

    International Nuclear Information System (INIS)

    Gravier, S.; Blandin, J.J.; Suery, M.

    2010-01-01

    A Metallic Glass/Alloy (MeGA) rod with a core in zirconium-based bulk metallic glass and a sleeve in aluminium alloy has been successfully elaborated by co-extrusion. SEM observations of the cross-section of the rod show that the interface between the glass and the alloy is defect-free. Compression tests are carried out at room temperature on the MeGA rods containing various glass volume fractions. The yield stress is well described by the rule of mixtures which combines the strength of the glass and that of the alloy, suggesting isostrain behaviour as could be expected. During compression, a good mechanical bonding is observed in the MeGA-rod even after the first fracture of the metallic glass. Finally, push-out tests are performed to evaluate the bonding quality between the two materials. Large values of the shear strength are measured which confirms that co-extrusion leads to good bonding between the glass and the aluminium alloy.

  18. Application of radiophotoluminescence dosemeters for environmental monitoring

    International Nuclear Information System (INIS)

    Mihara, A.; Tamura, T.; Mochizuki, T.; Numakunai, T.; Yoshida, Y.; Ohi, Y.

    1993-01-01

    Feasibility study has been made on the radiofluorescence glass (RPL) dosemeters for environmental radiation monitoring. Main items in the laboratory tests are batch uniformities of pre-dose and sensitivity, build-up characteristics of RPL intensity, energy and angular responses, stabilities for RPL reading, anneal cycling and climatic conditional changes. We have determined self-irradiation dose and cosmic ray contribution. Radiation monitoring field test for the RPL dosemeters has been conducted in comparison with the present TLD systems. A procedure manual for the environmental monitoring by the use of RPL dosemeters is in preparation

  19. Relative thermoluminescent efficiencies proton/gamma and helium/gamma of high temperature peaks in TLD-100 dosemeters; Eficiencias termoluminiscentes relativas proton/gamma y helio/gamma de picos de alta temperatura en dosimetros TLD-100

    Energy Technology Data Exchange (ETDEWEB)

    Flores M, E.; Avila, O.; Rodriguez V, M. [ININ, 52750 La Marquesa, Estado de Mexico (Mexico); Massillon, J.L.G.; Buenfil A, E.; Ruiz T, C.; Brandan, M.E. [IFUNAM, 04500 Mexico D.F. (Mexico); Gamboa De Buen, I. [ICN-UNAM, 04500 Mexico D.F. (Mexico)

    2007-07-01

    This work presents measures of relative thermoluminescent efficiency of those high temperature peaks of TLD-100 dosemeters exposed to protons of 1.5 MeV and to helium nuclei of 3 and 7.5 MeV. A rigorous reading and of deconvolution protocol was used for the calculation of the TL efficiencies. Additionally an Excel program that facilitated the deconvolution adjustment process of the glow curves was elaborated. (Author)

  20. Al{sub 2} O{sub 3}:Cr,Ni: a possible thermoluminescent dosemeter; Al{sub 2} O{sub 3}: Cr, Ni un posible dosimetro termoluminiscente

    Energy Technology Data Exchange (ETDEWEB)

    Mariani R, Francisco; Roman B, Alvaro; Saavedra S, Renato [Pontificia Univ. Catolica de Chile, Santiago (Chile). Facultad de Fisica; Ibarra S, Angel [Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas (CIEMAT), Madrid (Spain). Seccion Materiales para Fusion

    1997-12-31

    Results from a study on the thermoluminescent (Tl) emission from Al{sub 2} O{sub 3}:Cr,Ni are presented. The measurements were obtained for evaluation of the Al{sub 2} O{sub 3}:Cr,Ni dosimetric properties. Different crystal batches were exposed to two kind of ionizing radiation (X-ray and {beta}{sup -}). The Tl spectrum has a main peak with high thermal and optical stability, deviating from linearity for doses lower than 3.6 Gy. Furthermore, this material shows advantages (thermal resistance, reusability, multiple heating cycles) compared to TLD-100. Measured Al{sub 2} O{sub 3}:Cr,Ni properties indicate that it could be used as a dosemeter. (author). 5 refs., 4 figs.

  1. Re-assessment of dose from the Vinten extremity dosemeter

    International Nuclear Information System (INIS)

    O'Hagan, J.B.; Pearson, A.J.; Dutt, J.C.

    1989-01-01

    A procedure is described for re-assessing the dose from the Vinten extremity dosemeter using phototransferred thermoluminescence. The technique produces a linear response up to 50 Sv. The re-assessment efficiency is approximately 10% for the first re-assessment and 3-5% for the second re-assessment. The detection threshold values (at the 95% confidence level) are 3 mSv and 9 mSv respectively. (author)

  2. Calibration of personal dosemeters in terms of the ICRU operational quantities

    International Nuclear Information System (INIS)

    McDonald, J.C.; Hertel, N.E.

    1992-05-01

    The International Commission on Radiological Units and Measurements (ICRU) has defined several new operational quantities for radiation protection purposes. The quantities to be used for personal monitoring are defined at depths in the human body. Because these quantities are impossible to measure directly, the ICRU has recommended that personal dosemeters should be calibrated under simplified conditions on an appropriate phantom, such as the ICRU sphere. The US personal dosimetry accreditation programs make use of a 30 x 30 x 15 cm polymethymethacrylate (PMMA) phantom, therefore it is necessary to relate the response of dosemeters calibrated on this phantom to the ICRU operational quantities. Calculations of the conversion factors to compute dosemeter response in terms of the operational quantities have been performed using the code MCNP. These calculations have also been compared to experimental measurements using thermoluminescent (TLD) detectors

  3. Obtention of a thermoluminescent material for dosimetry of ionizing radiation; Obtencion de un material termoluminiscente para dosimetria de la radiacion ionizante

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez M, P R

    1990-07-01

    The thermoluminescent dosemeters are small crystals which suffer changes in their structure by the radiation effect, being displaced the electrons toward higher energy levels. On heating the previously irradiated crystals, the electrons come back to their base state emitting light photons. The light quantity emitted is proportional to the received radiation dose. The light quantity emitted is proportional to the received radiation dose. The lithium fluoride is one of the thermoluminescent materials considered as tissue equivalents by having a low effective atomic number (Z{sub ef}). At present, the more used commercial product used of this type is the TLD-100*. In this work the obtained results in the preparation of the lithium fluoride thermoluminescent material are presented. This is activated with magnesium (Mg) and titanium (Ti), which we have labelled as: LiF: Mg, Ti. The results from the tests performed for verifying his thermoluminescent properties are presented too, as powder form as in pellets form. These tests were performed in simultaneous form with TLD-100 samples, which is considered as reference. The LiF: Mg, Ti thermoluminescent material manufactured in the ININ presents similar dosimetric characteristics to those ones of the TLD-100. Therefore being able to replace the imported dosemeters. * (TLD-100 is a commercial trademark registered by Harshaw/Filtrol (US) for LiF: Mg, Ti Tl dosemeters) (Author)

  4. Physical properties and thermoluminescence of glasses designed for radiation dosimetry measurements

    International Nuclear Information System (INIS)

    Laopaiboon, R.; Bootjomchai, C.

    2015-01-01

    Highlights: • TL stability of soda-lime glass was corrected by dopants. • D LDL values indicated that the glass samples have good radiation sensitivity. • Bond compression model theory was used to confirm the results from experimental. • High elastic moduli of glass samples indicated that high stability of structure. - Abstract: Soda lime glasses doped with CeO 2 , Nd 2 O 3 and MnO 2 were prepared. Thermoluminescence (TL) properties, such as glow curves and linearity of TL response on irradiation dose were investigated. Results showed that the TL properties depended on the type and concentration of the dopants. Samples were selected to calculate energy trap depth parameters. To design materials for radiation dosimetry, physical properties, ion concentration, elastic properties and effective atomic numbers are important. Theoretical bond compression models were used to determine the elastic moduli for comparison with experimental values. Results show fair agreement between theoretical and experimental measurements. The high elastic moduli of the glass samples indicated high rigidity and stability of the glass matrix structure

  5. Photoluminescence and thermoluminescence properties of Li₂O-Na₂O-B₂O₃ glass.

    Science.gov (United States)

    Razak, N A; Hashim, S; Mhareb, M H A; Tamchek, N

    2016-05-01

    Influence of Nd(3+) concentration on the optical and thermoluminescence (TL) properties of melt-annealed synthesized 10 Na2O: 20 Li2O: (70-x) B2O3 : xNd2O3, where 0.1≤ x ≤0.7 (LNB) glasses are determined. The absence of sharp peaks in X-ray diffraction patterns confirms the amorphous nature of the prepared glasses. The photoluminescence spectra under 800 nm laser excitations at room temperature exhibit three prominent peaks centred at 538, 603 and 675 nm corresponding to the transitions of (4)G(7/2) → (4)I(9/2), [(4)G(7/2) → (4)I(11/2), (4)G(5/2) → (4)I(9/2)] and [(4)G(7/2) → (4)I(13/2), (4)G(5/2) → (4)I(11/2)], respectively. The TL glow curve exhibits a prominent peak (T(m)) at 180°C. The best performance of the prepared glass was found at 0.5 mol% of Nd2O3. We achieved a good linearity of TL response against dose between 0.5 to 4.0 Gy. The calculated value of the effective atomic number, Z(eff), is 7.55 which is nearly tissue equivalent (Z(eff) = 7.42). These promising features demonstrate the capability of the aforementioned glass to be used as a radiation dosimeter. The thermoluminescence and optical properties of new compositions of lithium sodium borate glasses doped with Nd(3+) ions were reported. Attractive features were obtained from the TL, PL and UV-Vis light analysis. Three upconversion luminescences permitting green, orange and red emissions were observed. Copyright © 2015 John Wiley & Sons, Ltd.

  6. Thermoluminescence dosimetry in Northwest Puerto Rico

    International Nuclear Information System (INIS)

    Block, A.M.; Clements, R.G.; Rosa, L.I.; Santos, F.

    1975-01-01

    In October, 1973, the Puerto Rico Water Resources Authority sought the initiation of studies aimed at determining background radiological characteristics of the northwestern quadrant of Puerto Rico using thermoluminescent dosemeters. Some of the studies were required for supporting data for the environmental report submitted as part of the licensing procedure in the establishment of thermonuclear electric power generation facilities in Barrio Islote, Arecibo, Puerto Rico. Previous studies of radiological characteristics of the area had been made using sodium iodide sensing equipment in an airplane flying at an altitude of 500 ft. The results are expressed in counts per second, and provide useful comparative levels for radioactivity measured with the thermoluminescent detectors. (U.S.)

  7. The IAEA/WHO thermoluminescent dosimetry intercomparison used for the improvement of clinical dosimetry

    International Nuclear Information System (INIS)

    Racoveanu, N.T.

    1981-01-01

    Results of thermoluminescent dosimetry collected over 5 years in the Eastern Mediterranean region of WHO were analyzed in an attempt to improve clinical dosimetry. Data for 16 radiotherapy departments showed considerable inconsistencies. It was found that the clinical dosemeters used by 3 of the departments were not working properly. The remainder of the departments had one or more dosemeters in perfect working order but the procedure for measuring machine output was inadequate or the correction factors (pressure, temperature) were wrongly applied due to lack of reliable instruments for such measurements. Problems encountered in the sending and returning of TLD dosemeters for assessment are discussed

  8. Thermoluminescent response of CaSO4: Dy + PTFE to beta particles

    International Nuclear Information System (INIS)

    Aguirre C, A.; Azorin N, J.

    2000-01-01

    In this work the results of studying the thermoluminescent properties of CaSO 4 : Dy + PTFE are presented when it is irradiated with beta particles. The conclusion was the obtention of the Tl response curve in function of dose is that to desexcite the dosemeters at temperature 300 C during 30 minutes and after that were irradiated at different times in groups and to do the reading of dosemeter, it can be observed that a greater irradiation time major is the Tl response and this depends of the material has been used. (Author)

  9. Exposure levels to radiation in a nuclear medicine laboratory: measurements with thermoluminescence dosemeters; Niveles de exposicion a la radiacion en un laboratorio de medicina nuclear: mediciones con dosimetros de termoluminiscencia

    Energy Technology Data Exchange (ETDEWEB)

    Ruiz J, A. [Departamento de Medicina Nuclear del Instituto nacional de Ciencias Medicas y Nutricion Salvador Zubiran, Vasco de Quiroga No. 15, Tlalpan 14000 Mexico D.F. (Mexico)

    2000-07-01

    Since 1996 in the Nuclear medicine Department and another routine and research departments of the National Institute of Medical and Nutrition Sciences (INCMNSZ) are working at least with 12 radioactive sources opened and sealed. However, it was unknown if with the Tl personal dosemeters with lithium fluoride crystals (LiF), could be possible to receive information about the equivalent dose for each radionuclide or if there was some radionuclide that by being low energy emissor or beta energy emissor, it did not represent an ionization power sufficient to excite the crystals of the thermoluminescent dosemeters (DTL). In this work the obtained results with control dosemeters of diverse source opened or sealed are shown, which were put at 1 cm of the LiF crystals during a time exposure 18 hours. Therefore the objective of this work is to verify which sources excite or not to DTL crystals and taking the pertinent safety measures for each radionuclide. The conclusion is that the majority of dosemeters were excited by beta or gamma radiation of the radionuclides and that LiF crystals are able to receive equivalent doses until 1200 mSv, being this a guarantee for the staff, that in an any moment could be exposed to high dose in his working day. Also it was corroborated that the radionuclides more energizer are: iodine-131, iodine-125 and sodium-22 while the phosphorus-32 must be managed with careful, since the exposure to hands can result significant, just like the technetium-99m ({sup 99m} Tc). (Author)

  10. Comparative study of thermoluminescent efficiencies before variation of experimental factors

    International Nuclear Information System (INIS)

    Avila, O.; Estrada, D.; Gonzalez, P.; Rodriguez V, M.; Buenfil, A.E.; Ruiz T, C.; Brandan, M.E.; Gamboa de Buen, I.; Aviles, P.; Horowitz, Y.S.

    2001-01-01

    An investigation for quantifying the influence of varying distinct experimental factors in the measurements of relative thermoluminescent efficiency of 3 MeV protons was realized with respect to gamma radiation in LiF: Mg, Ti. Variations in a dosemeters batch were considered, as presentation, applied thermal treatments and reader equipment. The main conclusion of this work is to emphasize the importance for measuring entire series of experiments in the same laboratory following a protocol defined carefully and using dosemeters of the same batch for obtaining response before charging particle and the response before gamma radiation with identical procedures of baking and reading. (Author)

  11. Development of confocal laser microscope system for examination of microscopic characteristics of radiophotoluminescence glass dosemeters

    International Nuclear Information System (INIS)

    Maki, D.; Ishii, T.; Sato, F.; Kato, Y.; Yamamoto, T.; Iida, T.

    2011-01-01

    A confocal laser microscope system was developed for the measurement of radiophotoluminescence (RPL) photons emitted from a minute alpha-ray-irradiated area in an RPL glass dosemeter. The system was composed mainly of an inverted-type microscope, an ultraviolet laser, an XY movable stage and photon-counting circuits. The photon-counting circuits were effective in the reduction of the background noise level in the measurement of RPL photons. The performance of this microscope system was examined by the observation of standard RPL glass samples irradiated using 241 Am alpha rays. The spatial resolution of this system was ∼3 μm, and with regard to the sensitivity of this system, a hit of more than four to five alpha rays in unit area produced enough amount of RPL photons to construct the image. (authors)

  12. Development of confocal laser microscope system for examination of microscopic characteristics of radiophotoluminescence glass dosemeters.

    Science.gov (United States)

    Maki, Daisuke; Ishii, Tetsuya; Sato, Fuminobu; Kato, Yushi; Yamamoto, Takayoshi; Iida, Toshiyuki

    2011-03-01

    A confocal laser microscope system was developed for the measurement of radiophotoluminescence (RPL) photons emitted from a minute alpha-ray-irradiated area in an RPL glass dosemeter. The system was composed mainly of an inverted-type microscope, an ultraviolet laser, an XY movable stage and photon-counting circuits. The photon-counting circuits were effective in the reduction of the background noise level in the measurement of RPL photons. The performance of this microscope system was examined by the observation of standard RPL glass samples irradiated using (241)Am alpha rays. The spatial resolution of this system was ∼ 3 μm, and with regard to the sensitivity of this system, a hit of more than four to five alpha rays in unit area produced enough amount of RPL photons to construct the image.

  13. How do hospital sterilisation procedures affect the response of personal extremity rings and of eye lens TL dosemeters?

    International Nuclear Information System (INIS)

    Kopec, Renata; Bubak, Anna; Budzanowski, Maciej; Sas-Bieniarz, Anna; Szumska, Agnieszka

    2016-01-01

    Stringent standards of hygiene must be applied in medical institutions, especially at operating blocks or during interventional radiology procedures. Medical equipment, including personal dosemeters that have to be worn by medical staff during such procedures, needs therefore to be sterilised. In this study, the effect of various sterilisation procedures has been tested on the dose response of extremity rings and of eye lens dosemeters in which thermoluminescent (TL) detectors (of types MTS-N and MCP-N, respectively) are used. The effects of medical sterilisation procedures were studied: by chemicals, by steam or by ultraviolet (UV), on the dose assessment by extremity rings and by eye lens dosemeters. Since it often happens that a dosemeter is accidentally machine-washed together with protective clothing, the effect of laundering on dose assessment by these dosemeters was also tested. The sterilisation by chemicals is mostly safe for TL detectors assuming that the dosemeters are waterproofed. Following sterilisation by water vapour, the response of these dosemeters diminished by some 30 %, irrespectively of the period of sterilisation; therefore, this method is not recommended. UV sterilisation can be applied to EYE-D TM eye lens dosemeters if their encapsulation is in black. The accidental dosemeter laundry in a washing machine has no impact on measured dose. (authors)

  14. Thermoluminescent relative efficiencies of TLD-100 for nitrogen ions respect of gamma radiation

    International Nuclear Information System (INIS)

    Concha S, K.; Avila, O.; Gamboa de Buen, I.; Rodriguez V, M.; Buenfil, A.E.; Ruiz T, C.; Brandan, M.E.

    2004-01-01

    The purpose of this work is to measure the thermoluminescent relative efficiency of those TLD-100 dosemeters irradiated with nitrogen ions with respect to the gamma radiation of 60 Co, in function of the linear energy transfer (LET). Two energy of such nitrogen ions were selected that has the same value from LET when impacting in the dosemeters but with E 1 energy (4.8 MeV) and E 2 (9.95 MeV) smaller and greater respectively that the energy of the Bragg peak. (Author)

  15. Investigation of dosimetric characteristics of the high sensitivity LiF:Mg,Cu,P thermoluminescent dosemeter and its applications in diagnostic radiology

    International Nuclear Information System (INIS)

    Fung, K.K.L.

    2000-12-01

    The recent introduction and development of the thermoluminescent (T.L.) phosphor material LiF:Mg,Cu,P (usually named TLD100H or GR200A) has aroused intense interest of scientists in the field of radiation dosimetry due to its very favourable dosimetric characteristics. Both conventional LiF:Mg,Ti (TLD100) and LiF:Mg,Cu,P T.L. phosphors are tissue-equivalent but GR200A outperforms in respect of its very much higher sensitivity, by a factor of greater than 25, and a dose detection threshold of less than 1 μGy. A reproducible readout and annealing regime was developed in the initial part of this study with the newly installed automatic TLD (Thermoluminescence Dosimetry) apparatus in the X-ray and Radiation Physics Laboratories of the Hong Kong Polytechnic University. Basic dosimetric characteristics of this T.L. dosemeter (supplied by Harshaw-Bicron Co.) were then investigated. This paved the foundation for subsequent selected novel application studies in diagnostic radiology. Dosimetric characteristics which included linearity, reproducibility, batch uniformity, energy response, and minimum detectable dose were studied using X-rays in the commonly used diagnostic radiology energy range. Favourable dosimetric characteristics were observed from this T.L. phosphor, which agrees well with published studies. The effect of the number of thermal treatment cycles in the initialisation process on dosimetric properties of this T.L. phosphor was also investigated. This study exploits the favourable dosimetric properties of these T.L. dosemeters in some selected novel dosimetric applications in diagnostic radiology with an anthropomorphic phantom using facilities both in these laboratories and also in radiology departments of various district hospitals in Hong Kong. Radiation absorbed dose from the direct or scattered beam, at critical sites inside and on the surface of the phantom, were measured in these radiological studies. The special focus in some of these studies was to

  16. Investigation of dosimetric characteristics of the high sensitivity LiF:Mg,Cu,P thermoluminescent dosemeter and its applications in diagnostic radiology

    Energy Technology Data Exchange (ETDEWEB)

    Fung, K.K.L

    2000-12-01

    The recent introduction and development of the thermoluminescent (T.L.) phosphor material LiF:Mg,Cu,P (usually named TLD100H or GR200A) has aroused intense interest of scientists in the field of radiation dosimetry due to its very favourable dosimetric characteristics. Both conventional LiF:Mg,Ti (TLD100) and LiF:Mg,Cu,P T.L. phosphors are tissue-equivalent but GR200A outperforms in respect of its very much higher sensitivity, by a factor of greater than 25, and a dose detection threshold of less than 1 {mu}Gy. A reproducible readout and annealing regime was developed in the initial part of this study with the newly installed automatic TLD (Thermoluminescence Dosimetry) apparatus in the X-ray and Radiation Physics Laboratories of the Hong Kong Polytechnic University. Basic dosimetric characteristics of this T.L. dosemeter (supplied by Harshaw-Bicron Co.) were then investigated. This paved the foundation for subsequent selected novel application studies in diagnostic radiology. Dosimetric characteristics which included linearity, reproducibility, batch uniformity, energy response, and minimum detectable dose were studied using X-rays in the commonly used diagnostic radiology energy range. Favourable dosimetric characteristics were observed from this T.L. phosphor, which agrees well with published studies. The effect of the number of thermal treatment cycles in the initialisation process on dosimetric properties of this T.L. phosphor was also investigated. This study exploits the favourable dosimetric properties of these T.L. dosemeters in some selected novel dosimetric applications in diagnostic radiology with an anthropomorphic phantom using facilities both in these laboratories and also in radiology departments of various district hospitals in Hong Kong. Radiation absorbed dose from the direct or scattered beam, at critical sites inside and on the surface of the phantom, were measured in these radiological studies. The special focus in some of these studies was to

  17. Study of a new glass matrix by the thermoluminescence technique; Estudo de uma nova matriz vitrea pela tecnica de termoluminescencia

    Energy Technology Data Exchange (ETDEWEB)

    Ferreira, Pamela Z.; Vedovato, Uly P.; Cunha, Diego M. da; Dantas, Noelio O.; Silva, Anielle C.A.; Neves, Lucio P.; Perini, Ana P., E-mail: anapaula.perini@ufu.br [Universidade Federal de Uberlandia (INFIS/UFU), MG (Brazil). Instituto de Fisica; Caldas, Linda V.E. [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Carrera, Betzabel N.S.; Watanabe, Shigueo [Universidade de Sao Paulo (IF/USP), Sao Paulo, SP (Brazil). Instituto de Fisica

    2015-07-01

    The thermoluminescence technique is widely used for both personal and for high-dose dosimetry. In this work, the thermoluminescence technique was utilized to study a new glass matrix, with nominal composition of 20Li{sub 2}CO{sub 3}.10Al{sub 2}O{sub 3}.20BaO.50B{sub 2}O{sub 3} (mol%), irradiated with different doses in a {sup 60}Co source. The glow curves and the dose-response curve were obtained for radiation doses between 50 Gy and 900 Gy. The results showed that this new glass matrix presents potential use in high-dose dosimetry. (author)

  18. Taylor impact of glass rods

    International Nuclear Information System (INIS)

    Willmott, G.R.; Radford, D.D.

    2005-01-01

    The deformation and fracture behavior of soda-lime and borosilicate glass rods was examined during classic and symmetric Taylor impact experiments for impact pressures to 4 and 10 GPa, respectively. High-speed photography and piezoresistive gauges were used to measure the failure front velocities in both glasses, and for impact pressures below ∼2 GPa the failure front velocity increases rapidly with increasing pressure. As the pressure was increased above ∼3 GPa, the failure front velocities asymptotically approached maximum values between the longitudinal and shear wave velocities of each material; at ∼4 GPa, the average failure front velocities were 4.7±0.5 and 4.6±0.5 mm μs -1 for the soda-lime and borosilicate specimens, respectively. The observed mechanism of failure in these experiments involved continuous pressure-dependent nucleation and growth of microcracks behind the incident wave. As the impact pressure was increased, there was a decrease in the time to failure. The density of cracks within the failed region was material dependent, with the more open-structured borosilicate glass showing a larger fracture density

  19. Thermoluminescent response of LiF before variation of the heating rate; Respuesta termoluminiscente de LiF ante variacion de la tasa de calentamiento

    Energy Technology Data Exchange (ETDEWEB)

    Barrios, R. [Facultad de Quimica, UAEM, 50000 Toluca, Estado de Mexico (Mexico); Avila, O. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    2003-07-01

    Comparisons of glow curves of lithium fluoride dosemeters TLD-100 measured to two heating rates with the purpose of quantifying the change in the temperature of the peaks 5 and 7 for the thermoluminescent reader equipment Harshaw 4000 of the thermoluminescence laboratory of the ININ were carried out. (Author)

  20. HOW DO HOSPITAL STERILISATION PROCEDURES AFFECT THE RESPONSE OF PERSONAL EXTREMITY RINGS AND OF EYE LENS TL DOSEMETERS?

    Science.gov (United States)

    Kopeć, Renata; Bubak, Anna; Budzanowski, Maciej; Sas-Bieniarz, Anna; Szumska, Agnieszka

    2016-09-01

    Stringent standards of hygiene must be applied in medical institutions, especially at operating blocks or during interventional radiology procedures. Medical equipment, including personal dosemeters that have to be worn by medical staff during such procedures, needs therefore to be sterilised. In this study, the effect of various sterilisation procedures has been tested on the dose response of extremity rings and of eye lens dosemeters in which thermoluminescent (TL) detectors (of types MTS-N and MCP-N, respectively) are used. The effects of medical sterilisation procedures were studied: by chemicals, by steam or by ultraviolet (UV), on the dose assessment by extremity rings and by eye lens dosemeters. Since it often happens that a dosemeter is accidentally machine-washed together with protective clothing, the effect of laundering on dose assessment by these dosemeters was also tested. The sterilisation by chemicals is mostly safe for TL detectors assuming that the dosemeters are waterproofed. Following sterilisation by water vapour, the response of these dosemeters diminished by some 30 %, irrespectively of the period of sterilisation; therefore, this method is not recommended. UV sterilisation can be applied to EYE-D™ eye lens dosemeters if their encapsulation is in black. The accidental dosemeter laundry in a washing machine has no impact on measured dose. © The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  1. Measurement of doses to aviator pilots using thermoluminescent dosemeters; Medicion de la dosis a pilotos aviadores usando dosimetros termoluminiscentes

    Energy Technology Data Exchange (ETDEWEB)

    Azorin N, J.; Cruz C, D. [UAM-I, 09340 Mexico D.F. (Mexico); Rivera M, T. [IPN, ESIME Culhuacan (Mexico)]. e-mail: azorin@xanum.uam.mx

    2004-07-01

    During the development of their work, the aviator pilots are exposed at high levels of natural radiation of bottom caused mainly by the cosmic radiation of galactic origin and lot. For such reason, the Metropolitan Autonomous University (UAM) and the Union Association of Aviator Pilots (ASPA), subscribed an agreement with the purpose of to measure the doses of ionizing radiation received by the aviator pilots of diverse air companies that man different types of airships and to determine if these doses surpass the one limit of 0.11 mSv/h settled down by the IAEA for the public in general; and if therefore, these workers should be considered as personnel occupationally exposed. In this work the obtained results when measuring the absorbed dose received by Mexican civil aviator pilots during the development of their work, using thermoluminescent dosemeters of LiF:Mg,Cu,P + Ptfe of national production are presented. The obtained results during the years of 2001 and 2002 show that the monthly doses received by the pilots surpass the one it limits established for the public in general, for what they should be considered as personnel occupationally exposed. (Author)

  2. Use of digital dosemeters for supporting staff radiation safety in paediatric interventional radiology suites

    International Nuclear Information System (INIS)

    McNeil, S. M.; Lai, P.; Connolly, B. L.; Gordon, C. L.

    2013-01-01

    Modern-day interventional radiology (IR) procedures impart a wide range of occupational radiation doses to team members. Unlike thermoluminescent badges, digital dosemeters provide real-time dose readings, making them ideal for identifying different components during IR procedures, which influence staff radiation safety. This study focused solely on paediatric IR (PIR) cases. Digital dosemeters measured the impact of imaging modality, shielding, patient and operator specific factors, on the radiation dose received during various simulated and real live PIR procedures. They recorded potential dose reductions of 10-to 100-fold to each staff member with appropriate use of shielding, choice of imaging method, staff position in the room and complex interplay of other factors. The digital dosemeters were well tolerated by staff. Results highlight some unique radiation safety challenges in PIR that arise from dose increases with magnification use and close proximity of staff to the X-ray beam. (authors)

  3. Use of digital dosemeters for supporting staff radiation safety in paediatric interventional radiology suites.

    Science.gov (United States)

    McNeil, Sarah M; Lai, Priscilla; Connolly, Bairbre L; Gordon, Christopher L

    2013-12-01

    Modern-day interventional radiology (IR) procedures impart a wide range of occupational radiation doses to team members. Unlike thermoluminescent badges, digital dosemeters provide real-time dose readings, making them ideal for identifying different components during IR procedures, which influence staff radiation safety. This study focused solely on paediatric IR (PIR) cases. Digital dosemeters measured the impact of imaging modality, shielding, patient and operator specific factors, on the radiation dose received during various simulated and real live PIR procedures. They recorded potential dose reductions of 10- to 100-fold to each staff member with appropriate use of shielding, choice of imaging method, staff position in the room and complex interplay of other factors. The digital dosemeters were well tolerated by staff. Results highlight some unique radiation safety challenges in PIR that arise from dose increases with magnification use and close proximity of staff to the X-ray beam.

  4. Accuracy and precision in thermoluminescence dosimetry

    International Nuclear Information System (INIS)

    Marshall, T.O.

    1984-01-01

    The question of accuracy and precision in thermoluminescent dosimetry, particularly in relation to lithium fluoride phosphor, is discussed. The more important sources of error, including those due to the detectors, the reader, annealing and dosemeter design, are identified and methods of reducing their effects on accuracy and precision to a minimum are given. Finally, the accuracy and precision achievable for three quite different applications are discussed, namely, for personal dosimetry, environmental monitoring and for the measurement of photon dose distributions in phantoms. (U.K.)

  5. Angular and radial dependence of the energy response factor for LiF-TLD micro-rods in 125I permanent implant source

    International Nuclear Information System (INIS)

    Mobit, P.; Badragan, I.

    2006-01-01

    EGSnrc Monte Carlo simulations were used to calculate the angular and radial dependence of the energy response factor for LiF-thermoluminescence dosemeters (TLDs) irradiated with a commercially available 125 I permanent brachytherapy source. The LiF-TLDs were modelled as cylindrical micro-rods of length 6 mm and with diameters of 1 mm and 5 mm. The results show that for a LiF-TLD micro-rod of 1 mm diameter, the energy response relative to 60 Co gamma rays is 1.406 ± 0.3% for a polar angle of 90 deg. and radial distance of 1.0 cm. When the diameter of the micro-rod is increased from 1 to 5 mm, the energy response decreases to 1.32 ± 0.3% at the same point. The variation with position of the energy response factor is not >5% in a 6 cm x 6 cm x 6 cm calculation grid for the 5 mm diameter micro-rod. The results show that there is a change in the photon spectrum with angle and radial distance, which causes the variation of the energy response. (authors)

  6. Study of some characteristic dosimetric of the LIF:Mg, You (JR1152C) for their employment as environmental dosemeter

    International Nuclear Information System (INIS)

    Molina Perez, D.; Diaz Bernal, E.; Prendes Alonso, M.

    1998-01-01

    As the interest grows in knowing the effects on the health of the drops dose of natural or artificial radiation, it is made but necessary a system dosimetric able to measure those dose levels accurately. P but of twenty years the dosemeters thermoluminescent (TL) they have constituted a simple and beautiful method for such mensurations. In the work the rehearsals of homogeneity, reproducibility, line lay, detection threshold, auto irradiation, residual, fading and angular dependence are described. The results demonstrate that the dosemeter satisfies the main requirements to be an employee in the environmental monitoring

  7. Thermoluminescence emission spectrometry of glass display in mobile phones and resulting evaluation of the dosimetric properties of a specific type of display glass

    International Nuclear Information System (INIS)

    Discher, Michael; Woda, Clemens

    2014-01-01

    Glass displays of mobile phones are sensitive to ionizing radiation and can be used for retrospective dosimetry for the purpose of triage after a radiological accident or attack. In this study the two main types of glass display that are used in modern mobile phones were investigated using thermoluminescence (TL) emission spectrometry. A different TL spectrum was observed for the glass display of category A (lime-aluminosilicate glass) and category B (boron-silicate glass). Based on the spectral measurements an optimized detection window was chosen to re-evaluate the dosimetric properties (dose response, optical and long-term stability) of glass display category B. - Highlights: • Two display glass types show similar TL emission peaks but with strongly different relative intensities. • The intrinsic background TL signal peaks at similar wavelengths as the radiation induced signal. • Dosimetric properties of one display glass type were re-evaluated using an optimized detection window

  8. Individual monitoring and its quality assurance using thermoluminescent CaSO[sub 4]:Dy Teflon tape dosemeters in India

    Energy Technology Data Exchange (ETDEWEB)

    Nagpal, J.S.; Popli, K.L.; Varadharajan, G.; Kher, R.K.; Venkataraman, G. (Bhabha Atomic Research Centre, Bombay (India). Div. of Radiological Protection)

    1994-01-01

    CaSO[sub 4]:Dy Teflon thermoluminscence dosemeters have been fabricated indigenously in continuous tape form. TL response of the tape dosemeters to [beta] and [gamma] radiations has been studied as a function of the dosemeter thickness to arrive at its optimum value. A single 80 mg.cm[sup -2] thick tape dosemeter (51 mm x 17 mm) along with a mica strip replaces the earlier use of three 200 mg.cm[sup -2] thick disc dosemeters for routine individual monitoring. The tape cards have been in field use for two and half years. The performance characteristics of the cards under laboratory and field conditions are discussed. Dose computation algorithms have been developed to give whole-body dose (mGy) and skin dose (mGy)for photons and [beta] radiation respectively from the TL outputs of three dosimetric elements under 1000 mg.cm[sup -2] Cu + Al composite filter, 180 mg.cm[sup -2] Perspex and an open region in a cardholder. Results of the Quality Assurance Programme, covering various TL monitoring stations in India, are also presented. Based on ANSI criteria, only one outlier was observed over the past seven years. (Author).

  9. N.R.P.B.'s new record-keeping service and thermoluminescent dosemeter

    International Nuclear Information System (INIS)

    Dennis, J.A.; Marshall, T.O.; Shaw, K.B.; Kendall, G.M.

    1977-01-01

    With the approval of the Health and Safety Executive the National Radiological Protection Board is introducing a radiation dose record keeping service in association with a thermoluminescent dosimeter. This service will relieve employers of the responsibility for maintaining and keeping dose records. Both the record keeping and dosimeter systems are described, including the documents which will be produced for employers. The thermoluminescent dosimeter is more accurate and less sensitive to environmental factors than the film dosimeter, which it replaces. It also reduces the consumption of materials and, with automated processing, is labour saving. (author)

  10. The development of a personal dosemeter for use by aircraft crew

    International Nuclear Information System (INIS)

    Stokes, R.P.; Talbot, L.

    2000-01-01

    This paper describes preliminary work to develop a cosmic-radiation dosemeter for use by military aircraft crew. The dosemeter is based on a combination of CR-39 etched-track detectors and TLD-700 thermoluminescent detectors. It is intended that the CR-39 be used to assess the neutron dose, while the TLD-700 is used to assess the photon and charged particle dose. The sensitivity of CR-39 to the neutron component of cosmic radiation was estimated by irradiating samples of the plastic at the CERN-CEC High Energy Reference Field Facility. This facility produced a radiation field with a neutron spectrum resembling that of the neutron component of cosmic radiation a typical airflight altitudes. The response of the CR-39 was linear over the range of doses studied (0.2-6.0 mSv) and there was no significant fading in the six-month period after irradiation. The TLD-700 component of the dosemeter was calibrated using 137 Cs gamma rays. The response of the TLD-700 was linear over the range of doses studied (0-1.1 mSv) with no significant fade in the six-month period after irradiation. It was concluded that a combination of CR-39 and TLD-700 detectors would provide an effective cosmic-radiation dosemeter for use by military aircraft crew. (author)

  11. A TLD reader for wide range in situ dose measurement

    International Nuclear Information System (INIS)

    Deme, S; Apathy, I.; Bodnar, L.; Csoeke, A.; Hejja, I.

    1998-01-01

    A portable thermoluminescent dosemeter reader suitable for reading at the site of exposure is described. The instrument consists of a glass bulb containing the CaSO 4 :Dy thermoluminescent material laminated to the surface of an electrically heated resistive metal plate. The bulb is encapsulated in a cylindrical, pen-like metal holder made of aluminium. A one-wire-port integrated electronic programmable memory chip mounted inside the holder contains the identification code and the individual calibration parameters of the dosemeter. The aperture of the holder is normally covered by a stainless steel tube to protect the bulb from light and mechanical shocks. The tube slips backwards automatically when the dosemeter is inserted into the reader. The system is used by the Hungarian nuclear power plant. (M.D.)

  12. Field calibration of a TLD albedo dosemeter in the high-energy neutron field of CERF

    International Nuclear Information System (INIS)

    Haninger, T.; Kleinau, P.; Haninger, S.

    2017-01-01

    The new albedo dosemeter-type AWST-TL-GD 04 has been calibrated in the CERF neutron field (CERN-EU high-energy Reference Field). This type of albedo dosemeter is based on thermoluminescent detectors (TLDs) and used by the individual monitoring service of the Helmholtz Zentrum Muenchen (AWST) since 2015 for monitoring persons, who are exposed occupationally against photon and neutron radiation. The motivation for this experiment was to gain a field specific neutron correction factor N n for workplaces at high-energy particle accelerators. N n is a dimensionless factor relative to a basic detector calibration with 137 Cs and is used to calculate the personal neutron dose in terms of H p (10) from the neutron albedo signal. The results show that the sensitivity of the albedo dosemeter for this specific neutron field is not significantly lower as for fast neutrons of a radionuclide source like 252 Cf. The neutron correction factor varies between 0.73 and 1.16 with a midrange value of 0.94. The albedo dosemeter is therefore appropriate to monitor persons, which are exposed at high-energy particle accelerators. (authors)

  13. Utilization of thermoluminescent dosemeters in radiation field studies for different materials

    International Nuclear Information System (INIS)

    Carron, W.; Guilardi Neto, T.

    1987-01-01

    The dose conversion factor of lithium fluoride thermoluminescent for the follows materials: water, wood and soil, the dose curve surveying in function of depth and the regression equations that better adapt to the dose curves were calculated. (C.G.C.) [pt

  14. Thermoluminescence of Zn3(PO4)2:Mn2+ exposed to β radiation

    International Nuclear Information System (INIS)

    Perez S, R.; Aceves, R.; Piters D, T.; Rodriguez M, R.; Camarillo, I.; Camarillo, E.

    2015-10-01

    The thermoluminescence (Tl) response and trapping analysis of Zn 3 (PO 4 ) 2 :Mn 2+ glass at room temperature after being exposed to a dose of β-irradiation is reported. The material shows a linear behavior for doses lower than 25 Gy. Measurements indicate that Tl glow curves are composed by several glow peaks. In order to verify this observation, partial heating technique has been applied. The result was the detection of at least four glow peaks located about 124, 199, 240, and 290 degrees C. The complete Tl glow curve shows a higher intensity at temperatures near 200 degrees C, which reduce possibilities of fading. A residual Tl analysis for different waiting times in the measurement of Tl indicate that low decay luminescent centers exist which correspond mainly to the first Tl glow peak. According with the result found, this type of material has potential to be used as β-irradiation dosemeter in different applications. (Author)

  15. Comparative measurements of external radiation exposure using mobile phones, dental ceramic, household salt and conventional personal dosemeters

    International Nuclear Information System (INIS)

    Ekendahl, Daniela; Bulánek, Boris; Judas, Libor

    2015-01-01

    Because retrospective dosimetry utilises commonly occurring materials and objects, it is particularly useful in cases of large-scale radiation accidents or malevolent acts with radioactive materials where casualties are inflicted on the general public and first emergency responders. The aim of this study was to investigate whether retrospective dosemeters can provide dose estimates with comparable accuracy like conventional personal dosemeters. Using an external source of radiation 137 Cs and an anthropomorphic phantom, we simulated serious irradiation of a human body in anterior-posterior and rotational geometries. Retrospective luminescence dosimetry objects, such as mobile phones, dental ceramic and household salt, and conventional personal dosemeters (thermoluminescent and electronic) were fixed to the anthropomorphic phantom. The doses obtained were compared with specific reference values. In most cases, relative deviations between the measured doses and the reference values did not exceed 20%. As the retrospective and conventional dosemeters show no significant differences in laboratory conditions, the retrospective luminescence dosimetry objects represent a very promising tool if handled properly. - Highlights: • A serious external exposure of human body was simulated. • Doses were measured using both retrospective and conventional dosemeters. • Utilised retrospective dosimetry materials were alumina resistors from mobile phones, household salt and dental ceramic. • Doses obtained were compared with reference values. • Both retrospective and conventional dosemeters gave similar results

  16. Thermoluminescence dating of some Hungarian medieval churches

    International Nuclear Information System (INIS)

    Kasa, I.; Bajnoczy, G.

    1984-01-01

    Thermoluminescence dating of three Hungarian historic churches was performed using the quarz inclusion technique and sup(60)Co gamma irradiation. Quarz grains obtained from the bricks were irradiated and the radiation doses were measured by CaSOsub(4):Dy TL dosemeters. Glow curves of irradiated and non-irradiated samples were also measured. From the results it was concluded that the ages of two churches were 10 and 30 percent less, respectively, than the ages estimated earlier. The age of the third church proved to be correct. (R.P.)

  17. Dosimetry of accidents using thermoluminescence of dental restorative porcelains

    International Nuclear Information System (INIS)

    Mauricio, C.L.P.; Rosa, L.A.R. da; Cunha, P.G. da

    1986-01-01

    The thermoluminescence (TL) properties of dental restorative porcelain were investigated with the aim of using this material as a TL dosemeter to estimate high doses in radiological accidents. The irradiations were carried out with a 60 Co gamma source and X rays with effective energies from 29 to 95 KeV. The samples have a limit of detection at about 50R and their reproducibility is better than 15%. Linearity was observed from 50 to 5000R. (Author) [pt

  18. Electroless silver coating of rod-like glass particles.

    Science.gov (United States)

    Moon, Jee Hyun; Kim, Kyung Hwan; Choi, Hyung Wook; Lee, Sang Wha; Park, Sang Joon

    2008-09-01

    An electroless silver coating of rod-like glass particles was performed and silver glass composite powders were prepared to impart electrical conductivity to these non-conducting glass particles. The low density Ag-coated glass particles may be utilized for manufacturing conducting inorganic materials for electromagnetic interference (EMI) shielding applications and the techniques for controlling the uniform thickness of silver coating can be employed in preparation of biosensor materials. For the surface pretreatment, Sn sensitization was performed and the coating powders were characterized by scanning electron microscopy (SEM), focused ion beam microscopy (FIB), and atomic force microscopy (AFM) along with the surface resistant measurements. In particular, the use of FIB technique for determining directly the Ag-coating thickness was very effective on obtaining the optimum conditions for coating. The surface sensitization and initial silver loading for electroless silver coating could be found and the uniform and smooth silver-coated layer with thickness of 46 nm was prepared at 2 mol/l of Sn and 20% silver loading.

  19. Measuring radiation exposure during percutaneous drainages: can shoulder dosemeters be used to estimate finger doses?

    International Nuclear Information System (INIS)

    Vehmas, Tapio; Tikkanen, Heikki

    1992-01-01

    To assess the need for extra finger dosemeters, radiologists' and assistants' radiation exposure at both shoulders and at the third fingers of both hands were recorded using thermoluminescent dosemeters during 27 interventional drainage procedures. Under couch screening was used. Mean dose rates were calculated by dividing dose by screening time. The radiologists' bilateral finger dose rates did not correlate with each other; nor did dose rates between the left shoulder and the right hand. The radiologists' dose rates at both shoulders, correlated with each other, as did shoulder dose rates with dose rates at the ispilateral hand. The right shoulder dose rates correlated with the left hand dose rates. The assistants' dose rates at places of measurement showed significant correlations with each other. (Author)

  20. Comparison of two models for the X-ray dispersion produced in a Novillo Tokamak with measurements make with thermoluminescent dosemeters; Comparacion de dos modelos para la dispersion de rayos X producidos en un Novillo Tokamak con mediciones efectuadas con dosimetros termoluminiscentes

    Energy Technology Data Exchange (ETDEWEB)

    Flores O, A.; Castillo, A.; Barocio, S.R.; Melendez L, L.; Chavez A, E.; Cruz C, G.J.; Lopez, R.; Olayo, M.G.; Gonzalez M, P. [Instituto Nacional de Investigaciones Nucleares, 52045 Salazar, Estado de Mexico (Mexico); Azorin N, J. [Universidad Autonoma Metropolitana Iztapalapa, 09340 Mexico D.F. (Mexico)

    1999-07-01

    It was presented the results to study about the X-ray dispersion produced in the Novillo Tokamak using thermoluminescent dosemeters (DTL). The measurements were make in the equatorial plane of Tokamak, along twelve radial directions. The dispersion is observed due to the radiation interaction with walls surrounding the machine. It was proposed two types of heuristic mathematical methods for describing the X-ray dispersion, comparing them with the experimental data obtained with Dtl. The predictions of both models are adjusted well to the experimental data. (Author)

  1. Superconducting glass-ceramics in BiSrCaCu2Al0.5Ox---Comparison between rod and powder compacted specimens

    International Nuclear Information System (INIS)

    Hirata, K.; Abe, Y.

    1991-01-01

    Superconducting properties are studied for glass-ceramics which were prepared by reheating glass rods and the glass powder compacts in the BiSrCaCu 2 Al 0.5 O x system, respectively. The glass-ceramic rod specimens obtained by reheating rod glass at 800--830 degree C for 50 h have a T c (R=0) of 85 K, while the disk specimens obtained by reheating the powered glass compacts in the same way do not exhibit superconductivity above 77 K. This difference in superconductivity between the specimens is discussed in terms of crystallization process and the amount of oxygen absorption of the specimens during heating

  2. Influence of alumina on photoluminescence and thermoluminescence characteristics of Gd{sup 3+} doped barium borophosphate glasses

    Energy Technology Data Exchange (ETDEWEB)

    Kalpana, T. [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, A.P. (India); Gandhi, Y. [Department of Physics, Kakani Venkata Ratnam College, Nandigama 521 185, A.P. (India); Sanyal, Bhaskar [Food Technology Division, Bhabha Atomic Research Center, Trombay, Mumbai 400 085 (India); Sudarsan, V. [Bhabha Atomic Research Centre, Chemistry Division, Mumbai 400 085 (India); Bragiel, P.; Piasecki, M. [Institute of Physics, Jan Dlugosz University, Ul. Armii Krajowej 13/15, Czestochowa 42-201 (Poland); Kumar, V. Ravi [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, A.P. (India); Veeraiah, N., E-mail: nvr8@rediffmail.com [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, A.P. (India)

    2016-11-15

    Gd{sup 3+} doped barium borophosphate glasses mixed with varying concentration of Al{sub 2}O{sub 3} are synthesized. Photoluminescence, thermoluminescence and other spectroscopic studies viz., IR and EPR spectral studies, have been carried out. IR spectral studies of these glasses indicated that there is a gradual increase in the degree of depolymerization of the glass network with increase in the concentration of Al{sub 2}O{sub 3} upto 3.0 mol%. The EPR spectral studies revealed the lowest concentration of Gd{sup 3+} ion clusters in the glass mixed with 3.0 mol% of Al{sub 2}O{sub 3}. The photoluminescence emission spectra exhibited strong ultraviolet blue emission at 311 under excitation at 273 nm due to {sup 6}P{sub 7/2}→{sup 8}S{sub 7/2} transition of Gd{sup 3+} ions. The intensity of this band is found to be enhanced four times when the glass mixed with 3.0 mol% of Al{sub 2}O{sub 3} with respect to that of alumina free glass. The enrichment of this emission is attributed to the declustering of Gd{sup 3+} ions by Al{sup 3+} ions. Thermoluminescence (TL) characteristics of these glasses have also been investigated after irradiating them with different doses of γ-rays (in the range 0–8.0 kGy). The TL emission exhibited a dosimetric peak at about 200 °C. The TL output under this glow peak is observed to increase with increase of γ-ray dose. For any fixed γ-ray dose, the TL output is increased with increasing Al{sub 2}O{sub 3} content up to 3.0 mol% and beyond this concentration quenching of TL is observed. The mechanisms responsible for TL emission and the variation in TL output with the concentration of Al{sub 2}O{sub 3} are quantitatively discussed in terms of electron and hole centers developed due to interaction of γ-rays with the glass network. The dose response of these glass samples exhibited linear behavior in the dose range 0–8.0 kGy.

  3. Investigations of the luminescence of phosphate glasses with respect to their application as solid dosemeters

    International Nuclear Information System (INIS)

    Regulla, D.F.

    1977-03-01

    A comprehensive presentation of the luminescent properties of phosphate glass is worked out. The extensive investigations show an essential enlargement of the knowledge on the luminescence of phosphate glass existing hitherto. These results form the basis for a detailed discussion of the luminescence mechanism. By applying additional results of measurements concerning optical and paramagnetic absorption, enabling the access to the atomic effect, propositions were developed for models of absorption, excitation and luminous centers relevant for dosimetry for which Ag ++ was found to be the constituent determining the centers. The interpretation of the luminescence phenomena within the frame of these models leads to considerable corrections on the existing concepts. At the same time the comparability of Ag-doped phosphate glasses and alkali halogenides is shown with respect to their luminescence behaviour, and with it an argument for the existence of crystal-like short-order regions in the amorphous glass is provided. This result serves as a basis for a discussion of the centers in the band model. Further investigations dealt with the quantities of influence for the practical application of the dosemeter. By interpretation of these results explanations are given for the effect of irradiation and evaluation temperatures, of LET, the dose, and UV light on the measuring signal. The phenomenon of 'pre-dose', for which especially surface effects have been detected as cause, is discussed under the aspect of a boundary layer theory. (orig./HP) [de

  4. Energy and angular dependence of the personal dosemeter in use at ITN-DPRSN

    Energy Technology Data Exchange (ETDEWEB)

    Alves, Joao G. [Departamento de Proteccao Radiologica e Seguranca Nuclear, Instituto Tecnologico e Nuclear, EN 10, 2686-953 Sacavem (Portugal)], E-mail: jgalves@itn.pt; Calado, Ana M.; Cardoso, Joao V.; Santos, Luis M. [Departamento de Proteccao Radiologica e Seguranca Nuclear, Instituto Tecnologico e Nuclear, EN 10, 2686-953 Sacavem (Portugal)

    2008-02-15

    In this paper the characterization of the dosimetry system and of the personal dosemeter in terms of the stability of the reader calibration factors and of the linearity of the response for the 137Cs reference radiation is presented. The energy and angular dependence of the whole body dosemeter are also shown. The energy dependence was determined performing irradiations with the X-ray narrow series beams N30, N40, N60, N80, N100, N120 and with the gamma reference radiations of 137Cs and 60Co [ISO 4037-1, 1996. X and Gamma Reference Radiation for Calibrating Dosemeters and Doserate Meters and for Determining Their Response as a Function of Photon Energy-Part 1: Radiation Characteristics and Production Methods. International Organization for Standardization, Geneva] in terms of Hp(10) incident on the ISO water slab phantom. The angular dependence of the dosemeter was determined for the angles 0 deg., {+-}20 deg., {+-}40 deg. and {+-}60 deg. with normal using the above mentioned radiation fields. All irradiations were performed at the Laboratorio de Metrologia das Radiacoes lonizantes of ITN-DPRSN. The experiments presented in this paper show the thermoluminescence dosimetry (TLD) system is stable and presents a linear behaviour over and extended dose range. The measurements allowed the determination of the energy dependence at normal incidence and of the angular dependence of the dosemeter currently in use. Further studies are being carried out in order to implement correction factors for supralinearity and low energy measurements.

  5. Calibration procedure for thermoluminescent dosemeters in water absorbed doses for Iridium-192 high dose rate sources

    International Nuclear Information System (INIS)

    Reyes Cac, Franky Eduardo

    2004-10-01

    Thermoluminescent dosimeters are used in brachytherapy services quality assurance programs, with the aim of guaranteeing the correct radiation dose supplied to cancer patients, as well as with the purpose of evaluating new clinical procedures. This work describes a methodology for thermoluminescent dosimeters calibration in terms of absorbed dose to water for 192 Ir high dose rate sources. The reference dose used is measured with an ionization chamber previously calibrated for 192 Ir energy quality, applying the methodology proposed by Toelli. This methodology aims to standardizing the procedure, in a similar form to that used for external radiotherapy. The work evolves the adaptation of the TRS-277 Code of the International Atomic Energy Agency, for small and big cavities, through the introduction for non-uniform experimental factor, for the absorbed dose in the neighborhood of small brachytherapy sources. In order to simulate a water medium around the source during the experimental work, an acrylic phantom was used. It guarantees the reproducibility of the ionization chamber and the thermoluminescent dosimeter's location in relation to the radiation source. The values obtained with the ionization chamber and the thermoluminescent dosimeters, exposed to a 192 Ir high dose rate source, were compared and correction factors for different source-detector distances were determined for the thermoluminescent dosimeters. A numeric function was generated relating the correction factors and the source-detector distance. These correction factors are in fact the thermoluminescent dosimeter calibration factors for the 192 Ir source considered. As a possible application of this calibration methodology for thermoluminescent dosimeters, a practical range of source-detector distances is proposed for quality control of 192 Ir high dose rate sources. (author)

  6. Development of an improved dosemeter for assessments of risk to the eye

    International Nuclear Information System (INIS)

    Eakins, J.S.; Gilvin, P.J.; Hager, L.G.; Tanner, R.J.

    2014-01-01

    To develop an improved dosemeter to assess photon and beta exposures of the eye lens, and in response to issues surrounding the preferred values of H lens to be used for guiding operational radiation protection, a programme of re-optimization of the current PHE thermoluminescence dosemeter has been performed. In particular, refinements of the filter located in front of the sensitive 7 LiF:Mg,Cu,P element have been considered, so that the dose response characteristics of the device provide a better and more conservative estimate of risk. The investigation was performed using the Monte Carlo modelling software MCNP5, to produce a final design that featured a filter containing a 9.5 mm diameter polypropylene hemisphere truncated to a maximum thickness of 3.0 mm. The responses of this design in photon and electron fields are presented here, contrasted against those of the existing PHE eye dosemeter, with respect to the operational quantity H p (3,E,θ) and both current and suggested values for the absorbed dose per fluence risk profile for the lens of the eye. - Highlights: • A programme of re-optimization of the current PHE eye dosemeter has been performed. • A design featuring a truncated hemispherical filter was found to be optimal. • The shape of the filter better resembles the rotational profile of the eye. • Response characteristics depend on the calibration conditions taken to provide the best risk estimate

  7. Thermoluminescent dose measurements on board Salyut type orbital stations

    International Nuclear Information System (INIS)

    Akatov, Yu.A.; Arkhangelskij, V.V.; Aleksandrov, A.P.

    1984-06-01

    A small, vibration- and shock-resistant thermoluminescent dosemeter (TLD) system - named PILLE - was developed for orbital stations at the Central Research Institute for Physics, Hungary, to measure the cosmic radiation dose on-board. The first on-board measurements with this system were performed by B. Farkas, the Hungarian astronaut, on the Salyut-6 space station in 1980. The same instrument was used by other crews in the following years. Doses measured at different sites in Salyut-6 are presented. The dose rates varied from 0.7 to 0.11 mGy.day -1 . After the first cosmic measurements, the system was further developed. The minimum detectable dose of the new TLD system is 1 μGy, i.e. less by one order of magnitude than that of the former system. The self-irradiation dose rate of the TLD bulbs is also reduced by more than an order of magnitude to 10 nGy.h -1 , by use of potassium-free glass for the bulb envelope. This new type of PILLE TLD system is currently on-board Salyut-7. The dose rates (0.12-0.23 mGy.day -1 ) measured in 1983 are presented in detail. (author)

  8. Thermoluminescence Response of Copper-Doped Potassium Borate Glass Subjected to 6 Megavolt X-Ray Irradiation

    Science.gov (United States)

    Hossain, I.; Shekaili, N. K.; Wagiran, H.

    2015-03-01

    This study addresses the characteristics of Cu-doped and undoped potassium borate glass for use as ionizing radiation dosimeters by investigating and comparing the thermoluminescence responses, linearity, sensitivity and dose response s of the two types of glasses. A number of samples based on xK 2 CO 3 + (100 - x)H 3 BO 3 , where 10 ≤ x ≤ 30 mol.%, have been prepared using a melt quenching technique. The amorphous phases were identified using X-ray diffraction (XRD). The undoped potassium borate samples 20K 2 CO 3 + 80H 3 BO 3 (mol.%) and Cu-doped (0.5 mol.%) samples were placed in a solid phantom apparatus and irradiated with in X-ray tube under 6 MV accelerating voltage with doses ranging from 0.5 to 4.0 Gy. This beam was produced by the Primus MLC 3339 linear accelerator (LINAC) available at Hospital Sultan Ismail, Johor Bahru, Malaysia. The results clearly show the superiority of Cu-doped glass in terms of response and sensitivity to producing luminescence over undoped potassium borate glass. The sensitivity of Cu-doped glass is 6.75 times greater than that of undoped glass.

  9. In vivo dosimetry using thermoluminescent detector in cancer therapy of head and neck

    International Nuclear Information System (INIS)

    Viegas, Claudio C.B.; Batista, D.V.; Campos, A.M.; Lopes, R.T.

    2002-01-01

    The viability and implementation of a routine in vivo dosimetry, using thermoluminescent dosemeters (TLD), at the radiotherapy section of the National Institute of Cancer in Brazil, in the case of head and neck treatment is shown. In order to reach that aim, the characteristics of the response of the LiF:Mg;Ti (TLD-100) thermoluminescent detectors in powder form were determined. The performed of this detector for in vivo dosimetry was testes using an RANDO Alderson anthropomorphic phantom and, once their adequability proved for the kind of measurements proposed , it was used for dose assessment in the case of tumour treatments in the head and neck regions, for Cobalt-60 irradiations. (author)

  10. Assessment of CVD diamond as a thermoluminescence dosemeter material

    International Nuclear Information System (INIS)

    Borchi, E.; Furetta, C.; Leroy, C.

    1996-01-01

    Diamond has a low atomic number (Z = 6) and is therefore essentially soft tissue (Z = 7.4) equivalent. As such, diamond is an attractive material for applications in dosimetry in which the radiation absorption in the sensor material should be as close as possible to that of soft tissue. Synthetic diamond prepared by chemical vapour deposition (CVD) offers an attractive option for this application. The aim of the present work is to report results on the thermoluminescence (TL) properties of CVD diamond samples. The annealing procedures, the linearity of the TL response as a function of dose, a short-term fading experiment and some kinetic properties have been investigated and are reported here. (Author)

  11. Technical recommendations for the use of thermoluminescence for dosimetry in individual monitoring for photons and electrons from external sources

    International Nuclear Information System (INIS)

    1975-01-01

    The principles of thermoluminescence dosimetry and the role it may play in the field of personnel dosimetry are presented. Technical recommendations are made as an indication of what is generally regarded as good practice and they are intended to be of assistance in avoiding the common sources of error. Errors due to: the detector; the evaluation procedure; the thermal treatment; at the sensitivity of the detector to neutrons. The recommendations are concerned with: the choice of detector material; design of thermoluminescent dosemeters; read-out procedure; reader; handling procedures; dose evaluations; automation; documentation and data handling

  12. Comparisons between direct ion storage and thermoluminescence dosimetry individual monitoring systems, and internet reporting

    International Nuclear Information System (INIS)

    Kiuru, A.; Kahilainen, J.; Hyvoenen, H.; Vartiainen, E.

    2001-01-01

    A new electronic direct ion storage (DIS) dosemeter allows accumulated personal dose equivalent H p (d) at depths of 10 mm and 0.07 mm to be monitored in a few seconds by inserting the dosemeter into a local reader without deleting the accumulated dose. The DIS system meets general requirements on individual monitoring of hospital personnel using ionising radiation. It differs greatly from off-line thermoluminescence dosimetry systems and offers many additional benefits. The non-volatile reading takes only 5 s, is taken as often as needed, and the data are collected into a dose database, where background radiation is subtracted. Individual personnel doses are reported in Intranet as well as on the Internet at regular intervals to the National Regulatory Authorities. (author)

  13. Thermoluminescence characteristics of Cu2O doped Calcium Lithium borate glass irradiated with the cobalt-60 gamma rays

    International Nuclear Information System (INIS)

    Rammadhan, Ismail; Taha, Saddon; Wagiran, H.

    2017-01-01

    The aim of this study is to prepare and investigate the thermoluminescence characteristics for the un-doped and Cu 2 O doped calcium lithium borate glass upon adding various Cu 2 O concentrations of 0.005% to 0.1 mol%. The glasses were prepared by melt quenching method and irradiated with 60 CO gamma-ray having different doses in the range of (0.5–4) Gy, (5–10) Gy, and (20–100) Gy. The amorphous phases were identified for optimization glass samples, effect of heating rate, glowing curves, linearity, sensitivity, fading, reproducibility of response and minimum detectable dose are also studied. The TL sample with 0.02 mol% Cu 2 O concentration has higher response compared to the other samples concentration for a delivered dose of 50 Gy, The recorded glow curves consist a dominant peak at 187 °C for a heating rate of 5 °C s −1 . However, the value of effective atomic number Z eff is 8.84 for 0.02Cu 2 O doped which are near to the atomic number of soft tissue. - Highlights: •We have prepared and investigate the crystalline structure for the un-doped and Cu 2 O doped calcium lithium borate glass are carried out. •Investigate the amorphous structure of calcium lithium borate glass. •Determine the best setting of annealing temperature, annealing time and heating rate for Cu 2 O doped calcium lithium borate glass. •The doping effects of Cu 2 O on the thermoluminescence properties of calcium lithium borate glass subjected gamma radiations. •The new dosimeter showed simple glow curve with single prominent peak centred at 187 °C and linear dose–response range 0.5–100 Gy, good reproducibility, the fading of the signal is relatively slow. •Effective atomic number for Cu 2 O doped Calcium lithium borate close to the effective atomic number of soft tissue.

  14. A review of environmental monitoring using solid state dosemeters, and guidelines for technical procedures

    International Nuclear Information System (INIS)

    Piesch, E.; Burgkhardt, B.

    1983-01-01

    The practical aspects of environmental programmes using thermoluminescence and photoluminescence dosemeter systems are reviewed. The review discusses developments in national and international standards, and discusses the optimisation of the read-out technique, routine calibration procedures and the treatment of random and systematic errors. The paper discusses the problems of measuring man-made dose contributions with low random errors, the estimation of this contribution being based on the interpretation of field results, taking into account the contribution of natural background radiation and of transit exposures. (author)

  15. Thermoluminescence of Zn{sub 3}(PO{sub 4}){sub 2}:Mn{sup 2+} exposed to β radiation

    Energy Technology Data Exchange (ETDEWEB)

    Perez S, R.; Aceves, R.; Piters D, T.; Rodriguez M, R. [Universidad de Sonora, Departamento de Investigacion en Fisica, Hermosillo, Sonora (Mexico); Camarillo, I. [Universidad Autonoma Metropolitana, Unidad Iztapalapa, Av. San Rafael Atlixco 186, Col. Vicentina, 09340 Mexico D. F. (Mexico); Camarillo, E., E-mail: rperez@cifus.uson.mx [UNAM, Instituto de Fisica, Cicuito Exterior, Ciudad Universitaria, 04510 Mexico D. F. (Mexico)

    2015-10-15

    The thermoluminescence (Tl) response and trapping analysis of Zn{sub 3}(PO{sub 4}){sub 2}:Mn{sup 2+} glass at room temperature after being exposed to a dose of β-irradiation is reported. The material shows a linear behavior for doses lower than 25 Gy. Measurements indicate that Tl glow curves are composed by several glow peaks. In order to verify this observation, partial heating technique has been applied. The result was the detection of at least four glow peaks located about 124, 199, 240, and 290 degrees C. The complete Tl glow curve shows a higher intensity at temperatures near 200 degrees C, which reduce possibilities of fading. A residual Tl analysis for different waiting times in the measurement of Tl indicate that low decay luminescent centers exist which correspond mainly to the first Tl glow peak. According with the result found, this type of material has potential to be used as β-irradiation dosemeter in different applications. (Author)

  16. Feasibility study of extremity dosemeter based on polyallyl-diglycol-carbonate (CR-39) for neutron exposure

    International Nuclear Information System (INIS)

    Chau, Q.; Bruguier, P.

    2007-01-01

    In nuclear facilities, some activities such as reprocessing, recycling and production of bare fuel rods expose the workers to mixed neutron-photon fields. For several workplaces, particularly in glove boxes, some workers expose their hands to mixed fields. The mastery of the photon extremity dosimetry is relatively good, whereas the neutron dosimetry still raises difficulties. In this context, the Inst. for Radiological Protection and Nuclear Safety (IRSN) has proposed a study on a passive neutron extremity dosemeter based on chemically etched CR-39 (PADC: polyallyl-diglycol-carbonate), named PN-3, already used in routine practice for whole body dosimetry. This dosemeter is a chip of plastic sensitive to recoil protons. The chemical etching process amplifies the size of the impact. The reading system for tracks counting is composed of a microscope, a video camera and an image analyser. This system is combined with the dose evaluation algorithm. The performance of the dosemeter PN-3 has been largely studied and proved by several laboratories in terms of passive individual neutron dosemeter which is used in routine production by different companies. This study focuses on the sensitivity of the extremity dosemeter, as well as its performance in the function of the level of the neutron energy. The dosemeter was exposed to monoenergetic neutron fields in laboratory conditions and to mixed fields in glove boxes at workplaces. (authors)

  17. Thermoluminescence kinetic analysis and dosimetry features of MgSO4:Dy and MgSO4:Cu nano-rods

    International Nuclear Information System (INIS)

    Zahedifar, M.; Almasifard, F.; Sadeghi, E.; Biroon, M. Kashefi; Ramazani- Moghaddam-Arani, A.

    2016-01-01

    MgSO 4 :Dy and MgSO 4 :Cu nano-rods (NRs) were synthesized for the first time by semi co- precipitation method. X-ray diffraction (XRD), scanning electron microscopy (SEM) and energy dispersive spectrometry (EDS) were utilized for sample characterization. The optimum amount of dysprosium and copper concentrations were obtained both at 0.1 mol% in MgSO 4 :Dy and MgSO 4 :Cu NRs. T m -T stop and computerized glow curve deconvolution (CGCD) methods were used for identifying the number of component glow peaks and kinetic parameters of the synthesized NRs. Initial rise and variable heating rate methods were also used to ensure the reliability of obtained kinetic parameters. Results show that the TL sensitivity of MgSO 4 :Dy is about 7 times more than that of magnesium sulfate doped with Cu. The TL dose response of MgSO 4 :Dy and MgSO 4 :Cu NRs are linear up to absorbed dose of 10 KGy. Other TL dosimetry characteristics of the produced NRs are also presented and discussed. - Highlights: • MgSO4:Dy and MgSO4:Cu nano-rods were synthesized for the first time. • Thermoluminescence dosimetry properties were studied. • The nano-phosphors showed linear dose response up to very high dose levels. • The synthesized nano-rods have potential application for high dose dosimetry.

  18. Glass transition of repulsive charged rods (fd-viruses).

    Science.gov (United States)

    Kang, Kyongok

    2014-05-14

    It has recently been shown that suspensions of long and thin charged fibrous viruses (fd) form a glass at low ionic strengths. The corresponding thick electric double layers give rise to long-ranged repulsive electrostatic interactions, which lead to caging and structural arrest at concentrations far above the isotropic-nematic coexistence region. Structural arrest and freezing of the orientational texture are found to occur at the same concentration. In addition, various types of orientational textures are equilibrated below the glass transition concentration, ranging from a chiral-nematic texture with a large pitch (of about 100 μm), an X-pattern, and a tightly packed domain texture, consisting of helical domains with a relatively small pitch (of about 10 μm) and twisted boundaries. The dynamics of both particles as well as the texture are discussed, below and above the glass transition. Dynamic light scattering correlation functions exhibit two dynamical modes, where the slow mode is attributed to the elasticity of helical domains. On approach of the glass-transition concentration, the slow mode increases in amplitude, while as the amplitudes of the fast and slow mode become equal at the glass transition. Finally, interesting features of the "transient" behaviors of charged fd-rod glass are shown as the initial caging due to structural arrest, the propagation of flow originating from stress release, and the transition to the final metastable glass state. In addition to the intensity correlation function, power spectra are presented as a function of the waiting time, at the zero-frequency limit that may access to the thermal anomalities in a charged system.

  19. Comparison test of electronic dosemeters

    International Nuclear Information System (INIS)

    Butterweck, G.; Zimmerli, H.P.; Wernli, C.

    2001-01-01

    To assist with a planned purchase of electronic dosemeters by the Swiss Federal Office for Civil Protection, the calibration laboratory of the Paul Scherrer Institute performed tests on 11 types of electronic dosemeters manufactured by 10 European and American companies. The technical specifications for the World Trade Organisation (WTO) tendering procedure were largely in accord with the specifications of the international standard IEC 61526. First tests were performed with samples from each type of dosemeter. The reproducibility of a dose of 0.1 mSv generated with 137 Cs radiation at a dose rate of 2.1 mSv.h -1 was found adequate for all tested dosemeter types. The response for environmental levels of radiation showed a large variation, indicating insufficient background correction of some dosemeters. A very high dose rate of 10 Sv.h -1 provoked faulty dose readings for more than half of the tested dosemeters. Dosemeter response for low-energy photon radiation was satisfactory for two of the tested dosemeter types. Four dosemeter types were selected for extended technical tests. Three samples of each of these dosemeter types were purchased. For drop and temperature tests the specifications of the WTO tendering procedure outranged the specification of the IEC standard. Whereas even at a temperature of -25 deg. C the tested dosemeters functioned normally, drops from a height of 2 m onto a wooden surface rendered the samples of two dosemeter types inoperative. (author)

  20. Thermoluminescence of magnesium doped zirconium oxide (ZrO2:Mg) UV irradiated

    International Nuclear Information System (INIS)

    Rivera Montalvo, Teodoro; Furetta, Claudio

    2008-01-01

    Full text: The monitoring of ultraviolet radiation (UVR) different thermoluminescent (TL) materials have been used to measure UVR. UV dosimetry using thermoluminescence phenomena has been suggested in the past by several authors. This technique has an advantage over others methods due to the readout of the samples. Other advantages of these phosphors are their small size, portability, lack of any power requirements, linear response to increasing radiation dose and high sensitivity. Zirconium oxide, recently received full attention in view of their possible use as thermoluminescent dosimeter (TLD), if doped with suitable activators, in radiation dosimetry. In the present investigation thermoluminescent (TL) properties of magnesium doped zirconium oxide (ZrO 2 :Mg) under ultraviolet radiation (UVR) were studied. The ZrO 2 :Mg powder of size 30-40 nm, having mono clinical structure, exhibit a thermoluminescent glow curve with one peak centered at 180 C degrees. The TL response of ZrO 2 :Mg as a function ultraviolet radiation exhibits four maxima centered at 230, 260, 310 and 350 nmn. TL response of ZrO 2 :Mg as a function of spectral irradiance of UV Light was linear in a wide range. Fading and reusability of the phosphor were also studied. The results showed that ZrO 2 :Mg nano powder has the potential to be used as a UV dosemeter in UVR dosimetry. (author)

  1. Thermoluminescence of glass display from mobile phones for retrospective and accident dosimetry

    International Nuclear Information System (INIS)

    Discher, M.; Woda, C.

    2013-01-01

    This paper deals with the thermoluminescence (TL) study of glass displays from mobile phones with the aim to use them as emergency dosimeters after an accident involving ionizing radiation. Dosimetric properties are analysed in order to examine and to critically evaluate the usability. Tests are carried out regarding the characterization of the radiation induced TL signal and the zero dose signal (intrinsic background) on a variety of display samples. Investigations on the thermal and optical stability of TL signals are carried out. The detection limit is mainly determined by the variability of the zero dose signal and lies in the range of 300–400 mGy. A linear relationship between the measured TL signal and the applied dose is observed for doses between 10 mGy and 20 Gy. A measurement protocol for the detection of absorbed radiation dose is developed, considering the experimental dosimetric properties. A reconstruction of the absorbed dose is possible using glass samples from mobile phones, if the signal loss due to storage and optical bleaching of the TL signal is adequately corrected for. This was confirmed by realistic tests. - Highlights: • Glass displays of mobile phones have potential for retrospective dosimetry. • Signal fading can be corrected with an universally fading curve. • Irradiation trials on intact mobile phones demonstrated a reasonable agreement between given and measured dose

  2. Acid dip for dosemeter

    International Nuclear Information System (INIS)

    Stewart, J.C.; McWhan, A.F.

    1982-01-01

    Background signal in a PTFE based dosemeter caused by impurities in the PTFE and in the active component such as lithium fluoride is substantially reduced by treating the dosemeter with acid. The optimum treatment involves use of hydrofluoric acid at room temperature for approximately one minute, followed by thorough washing with methanol, and finally drying. This treatment is best applied after the original manufacture of the dosemeters. It may also be applied to existing dosemeters after they have been in use for some time. The treatment produces a permanent effect in reducing both the light induced signal and the non-light induced signal. The process may be applied to all types of dosemeter manufactured from PTFE or other plastics or resins which are able to resist brief exposure to acid. The treatment works particularly well with dosemeters based on PTFE and lithium fluoride. It is also applicable to dosemeters based on calcium sulphate, lithium borate and magnesium borate. Acids which may be used include hydrofluoric, hydrochloric, nitric, phosphoric and sulphuric. (author)

  3. Thermoluminescence characteristics of Cu{sub 2}O doped Calcium Lithium borate glass irradiated with the cobalt-60 gamma rays

    Energy Technology Data Exchange (ETDEWEB)

    Rammadhan, Ismail, E-mail: ismail.rammadhan@koyauniversity.org [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Department of Physics, Faculty of Sciences and Health, Koya University, Danielle Mitterrand Boulevard, Koya 45, Kurdistan Region (Iraq); Taha, Saddon [Department of Physics, Faculty of Sciences and Health, Koya University, Danielle Mitterrand Boulevard, Koya 45, Kurdistan Region (Iraq); Wagiran, H. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia)

    2017-06-15

    The aim of this study is to prepare and investigate the thermoluminescence characteristics for the un-doped and Cu{sub 2}O doped calcium lithium borate glass upon adding various Cu{sub 2}O concentrations of 0.005% to 0.1 mol%. The glasses were prepared by melt quenching method and irradiated with {sup 60}CO gamma-ray having different doses in the range of (0.5–4) Gy, (5–10) Gy, and (20–100) Gy. The amorphous phases were identified for optimization glass samples, effect of heating rate, glowing curves, linearity, sensitivity, fading, reproducibility of response and minimum detectable dose are also studied. The TL sample with 0.02 mol% Cu{sub 2}O concentration has higher response compared to the other samples concentration for a delivered dose of 50 Gy, The recorded glow curves consist a dominant peak at 187 °C for a heating rate of 5 °C s{sup −1}. However, the value of effective atomic number Z{sub eff} is 8.84 for 0.02Cu{sub 2}O doped which are near to the atomic number of soft tissue. - Highlights: •We have prepared and investigate the crystalline structure for the un-doped and Cu{sub 2}O doped calcium lithium borate glass are carried out. •Investigate the amorphous structure of calcium lithium borate glass. •Determine the best setting of annealing temperature, annealing time and heating rate for Cu{sub 2}O doped calcium lithium borate glass. •The doping effects of Cu{sub 2}O on the thermoluminescence properties of calcium lithium borate glass subjected gamma radiations. •The new dosimeter showed simple glow curve with single prominent peak centred at 187 °C and linear dose–response range 0.5–100 Gy, good reproducibility, the fading of the signal is relatively slow. •Effective atomic number for Cu{sub 2}O doped Calcium lithium borate close to the effective atomic number of soft tissue.

  4. Thermoluminescent response of CaSO{sub 4}: Dy + PTFE to beta particles; Respuesta termoluminiscente de CaSO{sub 4}: Dy + PTFE a particulas beta

    Energy Technology Data Exchange (ETDEWEB)

    Aguirre C, A.; Azorin N, J. [Colegio de Bachilleres No. 13, Xochimilco-Tepepan, 16000 Mexico D.F. (Mexico)

    2000-07-01

    In this work the results of studying the thermoluminescent properties of CaSO{sub 4}: Dy + PTFE are presented when it is irradiated with beta particles. The conclusion was the obtention of the Tl response curve in function of dose is that to desexcite the dosemeters at temperature 300 C during 30 minutes and after that were irradiated at different times in groups and to do the reading of dosemeter, it can be observed that a greater irradiation time major is the Tl response and this depends of the material has been used. (Author)

  5. Solid thermoluminescent dosemeter of sodium tetraborate and brazilian fluoride sensitive to thermal neutrons

    International Nuclear Information System (INIS)

    Fratin, L.

    1988-01-01

    The techniques of compacting sodium tetraborate and natural fluoride mixtures were studied in this work, with the aim of producing a solid dosimeter sensitive to thermal neutrons. The production procedure involves the vitrification of the sodium tetraborate, the grinding, mixture, cold pressing and the sinterization of the pellets. A special arrangement was built for irradiation where paraffin was used as moderator for neutrons from a 241 Am-Be source. Two different mass ratios of sodium tetraborate and flourite showed a linear thermoluminescent response to the neutron fluence in the range of 1.0 to 7.0 x 10 8 n (sub)tcm -2 . Solid dosimeters, manufactured from natural fluorite and sodium chloride, showed a response to gamma radiation similar to the response of the dosimeters sensitive to neutrons. These dosimeters are need to identify the proportion of thermoluminescent response due to gamma radiation present in a neutron field. (author) [pt

  6. Thermoluminescence characterization of CVD diamond film exposed to UV and beta radiation

    International Nuclear Information System (INIS)

    Barboza-Flores, M.; Melendrez, R.; Gastelum, S.; Chernov, V.; Bernal, R.; Cruz-Vazquez, C.; Brown, F.; Pedroza-Montero, M.; Gan, B.; Ahn, J.; Zhang, Q.; Yoon, S.F.

    2003-01-01

    Thermoluminescence (TL) properties of diamond films grown by microwave and hot filament CVD techniques were studied. The main purpose of the present work was to characterize the thermoluminescence response of diamond films to ultraviolet and beta radiation. The thermoluminescence excitation spectrum exhibits maximum TL efficiency around 210-215 nm. All samples presented a glow curve composed of at least one TL peak and showed regions of linear as well as supralinear behavior as a function or irradiation dose. The linear dose dependence was found for up to sixteen minutes of monochromatic UV irradiation and 300 Gy for beta irradiated samples. The activation energy and the frequency factor were determined and found in the range of 0.33-1.7 eV and 5.44 x 10 2 -5.67 x 10 16 s -1 , respectively. The observed TL performance is reasonable appropriate to justify further investigation of diamond films as radiation dosimeters keeping in mind that diamond is an ideal TL dosemeter since it is tissue-equivalent and biological compatible. (copyright 2003 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  7. Values of dose and individual of a individual thermoluminescent dosimeter submitted to x and gamma radiations

    International Nuclear Information System (INIS)

    Moraes, Cassiana Viccari de; Pela, Carlos Alberto

    2001-01-01

    The individual monitoring provides information for the control of exposures, and estimates the dose received by individuals. This is an essential tool in personal dosimetry. It's based on a radiation protection concept, allowing an individual exposure control, besides guaranteeing that the dose restrictions will not be exceeded. Usually, the dose monitoring is performed by using an individual dosemeter placed on a representative position of the most exposed point on the thoracic surface. The dosemeter, which is analyzed in the present work, is made of three CaSO 4 -Dy thermoluminescent detectors, plastic filters, copper and copper-lead, mounted in an acrylic support. The dose received by on each detector, which forms the dosemeter, is related according to their energetic curve dependence. The dose amount is calculated from these curves by using an algorithm, and it was taken in to consideration the detector calibration and thermoluminescent responses, due to the x and g radiation exposure. That algorithm has the capacity to determine the energies that were irradiated the detector. Therefore, to aid the service in the moment of evaluate the dose received by the individual and where it is coming from. The algorithm has provided individual dose value H x , defined as operational quantity for photons adopted in the Brazilian Metric System. The algorithm can determine two dose values and such values have been analyzed according to the kind of irradiated energy on the dosimeter and it has shown that both values are within established limits by Instituto de Radioprotecao e Dosimetria (IRD). (author)

  8. Ultrasonic detection of cracks in uniaxial glass fibre rods

    CSIR Research Space (South Africa)

    Loveday, PW

    2006-01-01

    Full Text Available Conference on Computational and Applied Mechanics SACAM06 Cape Town, 16-18 January 2006 �SACAM ULTRASONIC DETECTION OF CRACKS IN UNIAXIAL GLASS FIBRE RODS Derren Wood and Philip Loveday Sensor Science and Technology, CSIR Materials Science... means of detecting internal and/or surface damage in composites which is safe, quick and relatively cost effective. Various ultrasonic techniques have been applied in the past to detect defects in composite media, the most well known being perhaps...

  9. Thermoluminescent characteristics of ZrO2:Nd films

    International Nuclear Information System (INIS)

    Vera B, G.; Rivera M, T.; Azorin N, J.; Falcony G, C.; Garcia H, M.; Martinez S, E.

    2002-01-01

    In this work it is exposed the obtained results after analysing the photo luminescent and thermoluminescent characteristics of activated zirconium oxide with neodymium (ZrO 2 :Nd) and its possible application in the UV radiation dosimetry. The realized experiments had as objective to study the characteristics such as the optimum thermal erased treatment, the influence of light on the response, the response depending on the wavelength, the fadeout of the information, the temperature effect, the response depending on the time and the recurring of the response. The results show that the ZrO 2 :Nd is a promising material to be used as Tl dosemeter for the UV radiation. (Author)

  10. Intrinsic dosimetry. Properties and mechanisms of thermoluminescence in commercial borosilicate glass

    Energy Technology Data Exchange (ETDEWEB)

    Clark, Richard A. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States)

    2012-10-01

    Intrinsic dosimetry is the method of measuring total absorbed dose received by the walls of a container holding radioactive material. By considering the total absorbed dose received by a container in tandem with the physical characteristics of the radioactive material housed within that container, this method has the potential to provide enhanced pathway information regarding the history of the container and its radioactive contents. The latest in a series of experiments designed to validate and demonstrate this newly developed tool are reported. Thermoluminescence (TL) dosimetry was used to measure dose effects on raw stock borosilicate container glass up to 70 days after gamma ray, x-ray, beta particle or ultraviolet irradiations at doses from 0.15 to 20 Gy. The TL glow curve when irradiated with 60Co was separated into five peaks: two relatively unstable peaks centered near 120 and 165°C, and three relatively stable peaks centered near 225, 285, and 360°C. Depending on the borosilicate glass source, the minimum measurable dose using this technique is 0.15-0.5 Gy, which is roughly equivalent to a 24 hr irradiation at 1 cm from a 50-165 ng source of 60Co. Differences in TL glow curve shape and intensity were observed for the glasses from different geographical origins. These differences can be explained by changes in the intensities of the five peaks. Electron paramagnetic resonance (EPR) and multivariate statistical methods were used to relate the TL intensity and peaks to electron/hole traps and compositional variations.

  11. Comparative study of thermoluminescent, radiophotoluminescent and photographic dosemeters

    International Nuclear Information System (INIS)

    Deus, S.F.

    1976-01-01

    Comparison was made between the response of three differents dosimetric systems, namely, photographic, thermoluminescent (TL) and radiophotoluminescent (RPL). The comparison was divided in two parts. The first was carried out in known radiation conditions (exposure, normal incidence, energy) and under controlled environment (approximately 27 0 C temperature, approximately 70% relative humidity). Under these conditions, the response as a function of exposure and energy, the relation of the linearity to the energy, the lowest detectable exposure, and the reproducibility, were studied. The response against esposure at 37 KeVef and at 1 MeV was found to be linear in the region of interest to routine personnel dosimetry for all dosimeters except for the filmes. In the second part, the relative response of the dosimeters was verified under the uncontrolled conditions of personnel dosimetry. As the CaSO 4 :Dy is the most sensitive dosimeter, comparison was made using this dosimeter as the standard, in which case one finds that 20 of 29 TLD-100 dosimeters give the same reading within 30%, 13 of 29 RPL dosimeters agree within 30%, and only 3 of 29 films fall within 30% [pt

  12. Dose evaluation in criticality accidents using response of panasonic TL personal dosemeters (UD-809/UD-802)

    International Nuclear Information System (INIS)

    Zeyrek, C. T.; Guenduez, H.

    2012-01-01

    This study gives the results of dosimetry measurements carried out in the Silene reactor at Valduc (France) with neutron and photon personal thermoluminescence dosemeters (TLDs) in mixed neutron and gamma radiation fields, in the frame of the international accident dosimetry intercomparison programme in 2002. The intercomparison consisted of a series of three irradiation scenarios. The scenarios took place at the Valduc site (France) by using the Silene experimental reactor. For neutron and photon dosimetry, Panasonic model UD-809 and UD-802 personal TLDs were used together. (authors)

  13. FT-IR and thermoluminescence investigation of P2O5-BaO-K2O glass system

    Science.gov (United States)

    Ivascu, C.; Timar-Gabor, A.; Cozar, O.

    2013-11-01

    The 0.5P2O5ṡxBaOṡ(0.5-x)K2O glass system (0≤x≤0.5mol%) is investigated by FT-IR and thermoluminescence as a possible dosimetic material. FT-IR spectra show structural network modifications with the composition variations of the studied glasses. The predominant absorption bands are characterized by two broad peaks near 500 cm-1, two weak peaks around 740 cm-1 and three peaks in the 900-1270 cm-1 region. The shift in the position of the band assigned to asymmetric stretching of PO2- group, υas(PO2-) modes from ˜1100 cm-1 to 1085 cm-1 and the decrease in its relative intensity with the increasing of K2O content shows a network modifier role of this oxide.. Luminescence investigations show that by adding modifier oxides in the phosphate glass a dose dependent TL signals result upon irradiation. Thus P2O5-BaO-K2O glass system is a possible candidate material for dosimetry in the dose 0 - 50 Gy range.

  14. Study about the feasibility of using an electret dosemeter as a personal dosemeter

    International Nuclear Information System (INIS)

    Leitao, I.P.

    1983-01-01

    A feasibility study of the electret dosemeters as a personnel dosemeter was carried out. Preliminary tests were applied, in order to assess the dosemeter response to radiation, linearity, stability, reproductibility, direcional dependence and energy dependence in energy range from 33 to 1250 KeV. In general the results are in agreement with the recommendations of the Commision of the European Communities (CEC) for the use of this device as a personnel dosemeter with the exception of the energy dependence and the stability at medium term. (Author) [pt

  15. Laser readable thermoluminescent radiation dosimeters and methods for producing thereof

    International Nuclear Information System (INIS)

    Braunlich, P.F.; Tetzlaff, W.

    1989-01-01

    Thin layer thermoluminescent radiation dosimeters for use in laser readable dosimetry systems, and methods of fabricating such thin layer dosimeters are disclosed. The thin layer thermoluminescent radiation dosimeters include a thin substrate made from glass or other inorganic materials capable of withstanding high temperatures and high heating rates. A thin layer of a thermoluminescent phosphor material is heat bonded to the substrate using an inorganic binder such as glass. The dosimeters can be mounted in frames and cases for ease in handling. Methods of the invention include mixing a suitable phosphor composition and binder, both being in particulate or granular form. The mixture is then deposited onto a substrate such as by using mask printing techniques. The dosimeters are thereafter heated to fuse and bond the binder and phosphor to the substrate. 34 figs

  16. A new electret dosemeter for radiotherapy

    International Nuclear Information System (INIS)

    Rodrigues, L.N.

    1985-01-01

    The results obtained with a new plane electret dosemeter for use in Radiotherapy are presented. The working principle of the electret dosemeter in based on the use of the electret field the ion collecting field, similarly to an ionization chamber, and as charge integrator detector. The dosemeter had, initialy, a cylindrical geometry. The results attained with it's calibration curve concerning the sensitivity, the stability and the reproducibility lead to changes in the dosemeter geometry. A plane geometry dosemeter was built in order to increase the useful exposure range. The plane dosemeter sensitivity dependence was measured versus the electret charging conditions, the chamber volume and wall material (aluminium, brass and stainless steel). The stainless steel dosemeter showed the best results. This dosemeter has the following characteristics: sensitivity 10 -13 C/R, volume 0.20 cm 3 . Compared to the Baldwin-Farmer dosemeter, this results are fairly good. A multiple electret dosemeter was developed to monitor the uniformity of the radiation beam for a Cobalt Teletherapy unit. The intercomparasion with the Baldwin-Farmer unit presented a maximum variation of 3%. (author) [pt

  17. Study of some thermoluminescent phosphors for the dosimetry of ionizing radiations

    International Nuclear Information System (INIS)

    Jaafari, M.

    1983-01-01

    Thermoluminescence dosimetry techniques are reviewed and interactions radiation matter are recalled. The need for new phosphors is evidenced. Numerous phosphors are examined and calcium, strontium and barium sulfides are synthetized and deposited on glass supports. The thermoluminescence of the dosimeters obtained with these materials is analyzed [fr

  18. Dose evaluation in criticality accidents using response of Panasonic TL personal dosemeters (UD-809/UD-802).

    Science.gov (United States)

    Zeyrek, C T; Gündüz, H

    2012-09-01

    This study gives the results of dosimetry measurements carried out in the Silène reactor at Valduc (France) with neutron and photon personal thermoluminescence dosemeters (TLDs) in mixed neutron and gamma radiation fields, in the frame of the international accident dosimetry intercomparison programme in 2002. The intercomparison consisted of a series of three irradiation scenarios. The scenarios took place at the Valduc site (France) by using the Silène experimental reactor. For neutron and photon dosimetry, Panasonic model UD-809 and UD-802 personal TLDs were used together.

  19. Thermoluminescent relative efficiencies of TLD-100 for nitrogen ions respect of gamma radiation; Eficiencias termoluminiscentes relativas de TLD-100 para iones de nitrogeno respecto de radiacion gamma

    Energy Technology Data Exchange (ETDEWEB)

    Concha S, K. [UNAM, Facultad de Ciencias, 04510 Mexico D.F. (Mexico); Avila, O. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico); Gamboa de Buen, I. [ICN-UNAM, 04510 Mexico D.F. (Mexico); Rodriguez V, M.; Buenfil, A.E.; Ruiz T, C.; Brandan, M.E. [IFUNAM, 04510 Mexico D.F. (Mexico)

    2004-07-01

    The purpose of this work is to measure the thermoluminescent relative efficiency of those TLD-100 dosemeters irradiated with nitrogen ions with respect to the gamma radiation of {sup 60} Co, in function of the linear energy transfer (LET). Two energy of such nitrogen ions were selected that has the same value from LET when impacting in the dosemeters but with E{sub 1} energy (4.8 MeV) and E{sub 2} (9.95 MeV) smaller and greater respectively that the energy of the Bragg peak. (Author)

  20. Machining a glass rod with a lathe-type electro-chemical discharge machine

    International Nuclear Information System (INIS)

    Furutani, Katsushi; Maeda, Hideaki

    2008-01-01

    This paper deals with the performance of electro-chemical discharge machining (ECDM) of a revolving glass rod. ECDM has been studied for machining insulating materials such as glass and ceramics. In conventional ECDM, an insulating workpiece is dipped in an electrolyte as a working fluid and a tool electrode is pressed on the surface with a small load. In the experiments, a workpiece was revolved to provide fresh working fluid into a gap between the tool electrode and the workpiece. A soda lime grass rod was machined with a thin tungsten rod in NaCl solution. The applied voltage was changed up to 40 V. The rotation speed was set to 0, 0.3, 3 and 30 min −1 . Discharge was observed over an applied voltage of 30 V. The width and depth of the machined grooves and the surface roughness of their bottom were increased with increase of the applied voltage. Although the depth of machining at 3 min −1 was the same as that at 30 min −1 , the width and roughness at 30 min −1 were smaller than those at 3 min −1 . Moreover, because the thickness of vaporization around the tool electrode was decreased with increase of the rotation speed, the width of the machined groove became smaller

  1. Relative thermoluminescent efficiencies proton/gamma and helium/gamma of peaks of high temperature in TLD-100 dosemeters; Eficiencias termoluminiscentes relativas proton/gamma y helio/gamma de picos de alta temperatura en dosimetros TLD-100

    Energy Technology Data Exchange (ETDEWEB)

    Flores M, E. [UNAM, Facultad de Ciencias, 04510 Mexico D.F. (Mexico)

    2007-07-01

    The increase of the applications of ion beams in radiotherapy treatments has generated interest in the study of the thermoluminescent materials (TL) that allow to determine the applied doses. A way to quantify the TL response from these materials to ions is by means of the relative thermoluminescent efficiency. In the group of Thermoluminescent dosimetry of the Institute of Physics of the UNAM (IFUNAM) the thermoluminescent response of the TLD-100 dosemeters has been studied, which present a glow curve characteristic with several peaks that correspond to traps and luminescent centers in the material. The stable peaks know each other as 4, 5, 6a, 6b, 7, 8, 9 and 10. The efficiencies should be measured using the response so much to the radiation of interest (in this case protons and helium ions) as the response to gamma radiation. In previous works with ions of low energy taken place in the Pelletron accelerator of the IFUNAM was only measured the TL efficiency for the peak 5 and the total signal. It had not been possible to measure the efficiency of the peaks of high temperature (6a-10) because, for the gamma radiation, the peaks of high temperature show very small signals; however, recently Massillon carries out measures of efficiency TL of peaks of high temperature for ions of intermediate energy using a protocol special of reading and of deconvolution that allows to measure the signals coming from the peaks of high temperature. In this work is implemented this same protocol to complete the study of TL efficiencies at low energy of protons and helium and to determine if the values of efficiency depend on the used reading protocol. For it is reported it measures of the relative efficiency of the peaks of high temperature from the TLD-100 exposed to protons of 1.5 MeV and nuclei of helium of 3 and 7.5 MeV. (Author)

  2. Thermoluminescent characterization of thin films of aluminium oxide submitted to beta and gamma radiation

    International Nuclear Information System (INIS)

    Villagran, E.; Escobar A, L.; Camps, E.; Gonzalez, P.R.; Martinez A, L.

    2002-01-01

    By mean of the laser ablation technique, thin films of aluminium oxide have been deposited on kapton substrates. These films present thermoluminescent response (Tl) when they are exposed to beta and gamma radiation. The brilliance curves show two peaks between 112 C and 180 C. A dose-response relationship study was realized and the Tl kinetic parameters were determined using the computerized deconvolution of the brilliance curve (CGCD). The thin films of aluminium oxide have potential applications as ultra.thin radiation dosemeters. (Author)

  3. Alarm personal dosemeter

    International Nuclear Information System (INIS)

    Nunomiya, Tomoya; Yamauchi, Hideshi; Shibata, Tetsuo

    2007-01-01

    Fuji Electric advanced the development of the conventional alarm personal dosemeter (APD) by adopting JIS Z 4312 (2002), which reflects IEC61526 (1998), and by improving the radiation performance characteristics, electromagnetic performance characteristics and mechanical performance characteristics of the APD (acquiring a CE marking). This new dosemeter will target overseas market in the future. The APD is a suitable dosemeter for detecting and monitoring radiation exposure to personnel working in the controlled areas of nuclear facilities, such as nuclear power stations. The APD product line includes 'gamma ray', 'gamma ray + beta ray' and 'gamma ray + neutron' models. (author)

  4. Evaluation of the thermoluminescent detector answers of CaSO4:Dy, LiF:Mg,Ti and micro LiF:Mg,Ti in photon clinical beams dosimetry using water simulator

    International Nuclear Information System (INIS)

    Matsushima, Luciana C.; Veneziani, Glauco R.; Campos, Leticia L.

    2011-01-01

    This paper perform the comparative study of thermoluminescent answer of calcium sulfate dosemeter doped with dysprosium (DaSO 4 :Dy) produced by the IPEN, Sao Paulo, with answer of lithium fluoride dosemeters doped with magnesium and titanium (LiF:Mg, Ti) in the dosimetry of clinical beams of photons (6 and 15 MV) by using water simulator object. Dose-answer curves were obtained for gamma radiation of cobalt-60 in the air and in conditions of electronic equilibrium (plate of PMMA), and clinical photons of CLINAC model 2100C accelerators of the two evaluated hospitals: Hospital das Clinicas of the Faculty of Medicine of Sao Paulo university and Hospital Albert Einstein. It was also evaluated the sensitivity and reproduction of the three dosemeters

  5. Report on a workshop on the application of thermoluminescence dosimetry to large scale individual monitoring, Ispra, 11-13 September 1985

    International Nuclear Information System (INIS)

    Barthe, J.R.; Christensen, P.; Driscoll, C.M.H.; Marshall, T.O.; Harvey, J.R.; Julius, H.W.; Marshall, M.; Oberhoffer, M.

    1987-01-01

    The workshop was held for the benefit of those actually involved in the operation of large scale automatic thermoluminescence dosimetry systems. It was organised around three overall themes subdivided into 13 subject headings: External constraints: User requirements, Quantities and Units, Legal requirements, Testing, Intercomparisons; Dosimetry systems: Materials, Dosemeter design, Reading systems, Annealing procedures, Rogue readings; Management: Distribution and organisation, Reporting and record keeping, Financial aspects. (author)

  6. Neutron dosimetry inside the containment building of Spanish nuclear power plants with PADC based dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Garcia-Fuste, M.J. [Grup de Fisica de les Radiacions. Departament de Fisica. Edifici C. Universitat Autonoma de Barcelona, E-08193 Bellaterra (Spain); Domingo, C., E-mail: carles.domingo@uab.ca [Grup de Fisica de les Radiacions. Departament de Fisica. Edifici C. Universitat Autonoma de Barcelona, E-08193 Bellaterra (Spain); Amgarou, K.; Bouassoule, T.; Castelo, J. [Grup de Fisica de les Radiacions. Departament de Fisica. Edifici C. Universitat Autonoma de Barcelona, E-08193 Bellaterra (Spain)

    2009-10-15

    The Spanish Nuclear Safety Council (Consejo de Seguridad Nuclear, CSN) recommends performing neutron individual dose assignments at workplaces based on ambient dose equivalent measurements using area monitors and by estimating the amount of time that workers spend in the different monitored environments. In addition, some Spanish nuclear power plants estimate the neutron dose equivalent using albedo thermoluminescence dosemeters (TLD). In the period 2004-2006, our group, together with other research centers, participated in a project, funded by the CSN, with the support of the Spanish Nuclear Power Plants Association (UNESA), to investigate in situ which could be the best practical procedure for individual neutron dose monitoring in nuclear power plants. As part of this survey, several units of the UAB PADC based neutron dosemeter were exposed, on a methacrylate phantom simulating a human body, at four different places inside the containment building of the Asco I nuclear power plant. The influence of different types of calibration neutron fields is analysed and the dose equivalent for each point is estimated.

  7. Analytical and regression models of glass rod drawing process

    Science.gov (United States)

    Alekseeva, L. B.

    2018-03-01

    The process of drawing glass rods (light guides) is being studied. The parameters of the process affecting the quality of the light guide have been determined. To solve the problem, mathematical models based on general equations of continuum mechanics are used. The conditions for the stable flow of the drawing process have been found, which are determined by the stability of the motion of the glass mass in the formation zone to small uncontrolled perturbations. The sensitivity of the formation zone to perturbations of the drawing speed and viscosity is estimated. Experimental models of the drawing process, based on the regression analysis methods, have been obtained. These models make it possible to customize a specific production process to obtain light guides of the required quality. They allow one to find the optimum combination of process parameters in the chosen area and to determine the required accuracy of maintaining them at a specified level.

  8. Design and test of the borosilicate glass burnable poison rod for Qinshan nuclear power plant core

    International Nuclear Information System (INIS)

    Huang Jinhua; Sun Hanhong

    1988-08-01

    Material for the burnable poison of Qinshan Nuclear Power Plant core is GG-17 borosilicate glass. The chemical composition and physico-chemical properties of GG-17 is very close to Pyrex-7740 glass used by Westinghouse. It is expected from the results of the experiments that the borosilicate glass burnable poison rod can be successfully used in Qinshan Nuclear Power Plant due to good physical, mechanical, corrosion-resistant and irradiaton properties for both GG-17 glass and cold-worked stainless steel cladding. Change of material for burnable poison from boron-bearing stainless steel to borosilicate glass will bring about much more economic benefit to Qinshan Naclear Power Plant

  9. Thermoluminescent response of LiF: Mg, Cu, P (GR-200) below an electron beam of 6 MeV

    International Nuclear Information System (INIS)

    Torijano C, E.F.S.; Azorin N, J.; Villasenor N, I.; Lujan C, P.J.; Rivera M, T.

    2007-01-01

    Full text: In this work the experimental results of studying the thermoluminescent response (TL) of LiF:Mg,Cu,P (GR-200) previously irradiated with 6 MeV electrons are presented. The electrons beam was generated by a Lineal Accelerator VARIAN I for medical use. The lineal accelerator is installed in the General Hospital of Mexico (HGM). A lot of 25 thermoluminescent dosemeters (DTL) was used. The mass and volume of each one of them were determined. Obtaining a variation of 14% in standard deviation (SD). The DTLs were irradiated to an energy of 6 MeV. The dose given to the DTL was of 50 c Gy. The linearity of the response of the GR-200 also was determined. (Author)

  10. Comparison measurements of the official individual dosimetry systems in the Federal Republic of Germany

    International Nuclear Information System (INIS)

    Boehm, J.

    1988-01-01

    At present, about 220,000 radiation workers are subject to individual monitoring of external radiation in the Federal Republic of Germany in accordance with the X-Ray Ordinance and the Radiation Protection Ordinance and the Radiation Protection Ordinance. Each radiation worker must wear at least one individual dosemeter provided by one of the five officially authorized dosemeter services. Requirements for these dosemeter services with respect to technical equipment and personnel are published in the form of a national directive. The five dosemeter services must participate in the comparison measurements for photon radiation with the official dosemeters issued by them. The comparison measurements are organized at least once a year by the Physicalisch-Technische Bundesanstalt (PTB) in agreement with the Federal Minister of Economics. Film dosemeters evaluated with a filter analysis procedure are most frequently used (96.% of all evaluations in 1984) to monitor the whole body exposure to photon radiation. Glass dosemeters play a minor role (0.5%). They are issued by only two officially authorized dosemeter services and are approved for measurements of photons with energies above 45 keV. According to the national directive, thermoluminescence dosemeters may be used only for partial body dosimetry of photon and beta radiation (2.8%). Nuclear track emulsion films are distributed to measure neutrons of energies above 0.5 MeV. A thermoluminescence albedo dosemeter for the measurement of less energetic neutrons is being introduced. The dosimetry systems involved in the comparison measurements and the operational radiation protection monitoring system of the Federal Republic of Germany are described elsewhere. The objective of this work is to describe the performance of the comparison measurements and to give some results obtained from 1979 to 1984

  11. Experimental calibration and determination of the relative response for Lif: Mg, Ti(TLD-100) dosemeters at 60Co gamma and 60 kVp X-ray energies

    International Nuclear Information System (INIS)

    Marco, M.L.; Gonzalez, L.; Delgado, V.; Vano, E.; Moran, P.

    1985-01-01

    The thermoluminescence efficiency of LiF: Mg, Ti (TLD-100) dosemeters has been determined for photon beams from 60 Co gamma rays and 60 kVp X-rays. It has been proven that light yield varies as a function of the photon energy. An experiment was performed using an X-ray beam whose spectrum has been determined by an X-ray fluorescence method. This enabled a direct calculation of the absorbed doses in the T1 material for the different operation conditions. These values and the experimental ones from measuring T1 intensities have been used to obtain the efficiency for energy X-ray spectrum. From the above values, the dosemeter T1 response, relative to 60 Co, has been evaluated. (author)

  12. Investigation about the thermal features of the ovens used for thermoluminescence

    International Nuclear Information System (INIS)

    Scarpa, G.; Caporali, C.; Moscati, M.

    1991-02-01

    The present paper reports the results of an investigation carried out by the PAS-FIBI-DOSIBIO laboratory (ENEA, Casaccia, Roma) about the thermal features of the ovens used for annealing treatments of TL dosemeters. A total number of 45 commercial ovens and muffle furnaces were studied. belonging to 24 Italian Health Physics laboratories. The investigation has shown that the majority of the ovens do not possess a degree of accuracy, stability, uniformity and reproducibility suitable for their use in the field of thermoluminescence dosimetry. Practical suggestions are also given in order to reduce the effects of some of the negative characteristics found in most ovens. (author)

  13. Thermoluminescence: Potential Applications in Forensic Science

    Science.gov (United States)

    Ingham, J. D.; Lawson, D. D.

    1973-01-01

    In crime laboratories one of the most difficult operations is to determine unequivocally whether or not two samples of evidence of the same type were originally part of the same thing or were from the same source. It has been found that high temperature thermoluminescence (room temperature to 723 K) can be used for comparisons of this type, although work to date indicates that there is generally a finite probability for coincidental matching of glass or soil samples. Further work is required to determine and attempt to minimize these probabilities for different types of materials, and to define more clearly the scope of applicability of thermoluminescence to actual forensic situations.

  14. Fast neutron sensitivity of polymer dosemeters

    International Nuclear Information System (INIS)

    Harper, M.W.; Pearson, D.W.; Moran, P.R.

    1975-01-01

    The responses of polymer thermocurrent dosemeters to fission spectrum and 14 MeV neutrons were measured. The dosemeters are in the form of disks 1 cm diam by 0.5 mm thick. Relative to Cobalt 60 gamma responses, teflon PTFE dosemeters show a 6 percent response to 14 MeV neutrons and a 5 percent response to fission neutrons on a tissue rad basis. Polymethylpentene dosemeters show a 49 percent response to 14 MeV neutrons and a 40 percent response to fission neutrons on a tissue rad basis when provided with adequate recoil proton buildup. The sensitivity of these dosemeters is limited to neutron doses greater than 10 rads by spurious background currents

  15. Quality control for dosemeters

    International Nuclear Information System (INIS)

    Di Prinzio, R.

    1988-01-01

    Several tests and intercomparison of the secondary standard dosemeter calibration in Rio de Janeiro are described. Periodical tests, such as leakage current and the comparison between the secondary standard dosemeter in Rio de Janeiro e Sao Paulo are shown. (E.G.) [pt

  16. Development of Mg B4 O7:Dy thermoluminescent solid dosemeter for using on the thermal neutron dosimetry

    International Nuclear Information System (INIS)

    Potiens Junior, A.J.; Campos, L.L.; Todo, A.S.

    1992-01-01

    The magnesium tetraborate doped with dysprosium (Mg B 4 O 7 :O y ) thermoluminescent phosphor is developed, containing boron, which boron 10 isotope has high cross section for thermal neutrons by 10 B (n,α) 7 Li reaction. The nitric acid concentration, the number of washing for removing the excess of boric acid, the powder granulation of Mg B 4 O 7 :O y , the percentage of teflon used as binder and the reproducibility of thermoluminescent answer are also presented. (C.G.C.)

  17. Personal neutron diode dosemeter

    International Nuclear Information System (INIS)

    Barthe, J.; Lahaye, T.; Moiseev, T.; Portal, G.

    1993-01-01

    The control and management of neutron doses, received by workers in nuclear power or research facilities, requires a knowledge of cumulated dose equivalent or dose equivalent rate in real time. Individual dosemeters so far developed for this purpose are scarce and not very satisfactory. Passive dosemeters such as TLD systems based on the albedo effect, nuclear emulsions or solid track detectors, do not give sufficiently accurate measurements. Furthermore, the increase in the quality factor and the more restrictive new ICRP recommendations diminish the maximum admissible threshold making currently used systems obsolete. Other than bubble dosemeter systems, based on thermodynamic effects of a superheated gel, no simple electronic device is available at the present time. The development of diode based dosimetric gamma badges, having a size similar to that of credit cards, has stimulated us to design and develop a personal neutron dosemeter based on a double diode system. The results obtained are very encouraging and practical models should become available in the near future. (author)

  18. Some studies about the NaCl:Ca2+ :Mn2+ and NaCl: Cd2+ :Mn2+ dosemeters

    International Nuclear Information System (INIS)

    Verdiguel G, H.; Flores J, C.; Camarillo G, E.; Espejel P, R.; Cabrera B, E.; Hernandez A, J.; Murrieta S, H.; Cruz Z, E.; Ramos B, S.; Negron, A.

    2002-01-01

    Nowadays, a great interest by counting with dosemeters of characteristics such as a high stability, of easy operation and easier production exists. Looking for a commitment with all these characteristics,a possibility to use the system NaCl: Ca 2+ :Mn 2+ and NaCl: Cd 2+ :Mn 2+ as dosemeters was studied. The studies were realized irradiating with gamma radiation from a 60 Co source. The crystals that were used as samples did not suffer any thermal treatment previous to irradiation. The supplied doses were 10, 30, 60, 100, 300, and 600 rads. 24 hours after irradiation the thermoluminescent response was obtained. In the case of the system NaCl: Ca 2+ :Mn 2+ several thermoluminescent bands were observed (BTL). Two concentrations of Mn 2+ with only one concentration of Ca 2+ (1%) were studied. For the case of the smaller concentration of Mn 2+ (0.1%) 4 BTL were observed, whereas for a greater concentration (0.3%) just 2 BTL were detected. The positions of the maximum of the BTL peaks differ for both concentrations, this possible due to what the nature of the traps for both cases differs by the type of precipitates present in the net. For the case of the system NaCl: Cd 2+ (1%) :Mn 2+ (0.1% and 0.5%) a similar situation to the previous was found, although in this case for both manganese concentrations just 2 BTL were observed; however all the peaks seem to be the superposition of several bands. Despite the apparent complexity of the thermoluminescent response, such response as function of the dose shows that both systems present a stable response to gamma radiation in the interval from 10 to 600 rads. In the case of calcium it is had a response of linear type of the Tl intensity depending on the dose, whereas for the cadmium system a supra linear response seems to exist. Nowadays, studies for determining the BTL origin being carried out. (Author)

  19. Glass badge dosimetry system for large scale personal monitoring

    International Nuclear Information System (INIS)

    Norimichi Juto

    2002-01-01

    Glass Badge using silver activated phosphate glass dosemeter was specially developed for large scale personal monitoring. And dosimetry systems such as an automatic leader and a dose equipment calculation algorithm were developed at once to achieve reasonable personal monitoring. In large scale personal monitoring, both of precision for dosimetry and confidence for lot of personal data handling become very important. The silver activated phosphate glass dosemeter has basically excellent characteristics for dosimetry such as homogeneous and stable sensitivity, negligible fading and so on. Glass Badge was designed to measure 10 keV - 10 MeV range of photon. 300 keV - 3 MeV range of beta, and 0.025 eV - 15 MeV range of neutron by included SSNTD. And developed Glass Badge dosimetry system has not only these basic characteristics but also lot of features to keep good precision for dosimetry and data handling. In this presentation, features of Glass Badge dosimetry systems and examples for practical personal monitoring systems will be presented. (Author)

  20. Thermoluminescent dosimetry in dentistry students in radiological training

    International Nuclear Information System (INIS)

    Loya, M.; Duarte, R.; Montero, M.E.; Gonzalez, P.R.; Ojeda, S.L.; Sanin, L.H.

    2007-01-01

    Full text: In this work, the obtained results of the mensuration of the equivalent one are presented of dose (E), in students of the career of Dentistry who during their training in radiology complete the three lists of patient, odontologist and observer. The study one carries out with 35 students (Odontologists), 7 men and 28 women, with ages between 20 and 24 years. A characterization of risk was made following the methodology of the Agency for the environmental control of United States (EPA), counting the number of shots in each list, during the time of training. The effective energy of the X rays used was of 24 KeV. The irradiation technique understood two modalities; the first one with 3 shots of 1.25 seconds and the second with 5 shots of 0.6 seconds. The measurement of E, was carried out with thermoluminescent dosemeters (TL) of LiF:Mg,Cu,P+PTFE, developed in the National Institute of Nuclear Research (ININ). The homogenization of the dosemeters showed a variation in its response to the radiation of 1.85%DS. For the odontologist position, the dosemeters was placed to the height of the chest and the measurement was multiplied by the number of shots that it was 304. The E value was of 3.62 mSv/four-month (11 mSv/year), without any safety equipment and of 2.02 mSv/four-month (6.14 mSv/year), when was considered the scenario for this position with use of vest and lead collar, like safety equipment, diminishing E, almost 50%. In the case of the corresponding position to patient, E, it was measured in different organs of interest, the obtained results are compared with the international standards. Recommendations are given for similar cases. (Author)

  1. Self-indicating radiation alert dosemeter (SIRAD)

    International Nuclear Information System (INIS)

    Riel, G. K.; Winters, P.; Patel, G.; Patel, P.

    2006-01-01

    In an event of a nuclear or dirty bomb explosion and a radiological accident, there is a need for self-indicating instant radiation dosemeter for monitoring radiation exposure. The self-indicating instant radiation alert dosemeter (SIRAD) is a credit card size radiation dosemeter for monitoring ionising radiation from a few hundredths of a Gray to a few Gray. It is always active and is ready to use. It needs no battery. The dosemeter develops colour instantly upon exposure, and the colour intensifies with dose. It has a colour chart so that the dose on the active element may be read by matching its colour with the chart that is printed next to it on the card. However, in this work, the dose is measured by the optical density of the element. The dosemeter cannot be reset. The response changes by 3 y at room temperature. It contains no hazardous materials. The dosemeter would meet the requirements of instantly monitoring high dose in an event of a nuclear or dirty bomb explosion or a radiation accident. (authors)

  2. Self-indicating radiation alert dosemeter (SIRAD).

    Science.gov (United States)

    Riel, Gordon K; Winters, Patrick; Patel, Gordhan; Patel, Paresh

    2006-01-01

    In an event of a nuclear or dirty bomb explosion and a radiological accident, there is a need for self-indicating instant radiation dosemeter for monitoring radiation exposure. The self-indicating instant radiation alert dosemeter (SIRAD) is a credit card size radiation dosemeter for monitoring ionising radiation from a few hundredths of a Gray to a few Gray. It is always active and is ready to use. It needs no battery. The dosemeter develops colour instantly upon exposure, and the colour intensifies with dose. It has a colour chart so that the dose on the active element may be read by matching its colour with the chart that is printed next to it on the card. However, in this work, the dose is measured by the optical density of the element. The dosemeter cannot be reset. The response changes by 3 y at room temperature. It contains no hazardous materials. The dosemeter would meet the requirements of instantly monitoring high dose in an event of a nuclear or dirty bomb explosion or a radiation accident.

  3. Oxidation/reduction effects on sapphire thermoluminescent detectors

    International Nuclear Information System (INIS)

    Molnar, G.; Borossay, J.; Suevegh, K.; Vertes, A.

    1998-01-01

    Complete text of publication follows. The thermoluminescence (TL) and thermally stimulated exoemission (TSEE) properties of Al 2 O 3 has been extensively studied during the past 40 years. It has been suggested as a dosemeter for both ultraviolet light and ionising radiation. Nevertheless the connection between the structure (point defects) of alumina and thermally stimulated relaxation processes has not yet been clearly explained. In this work α-Al 2 O 3 single crystal samples were studied in order to determine the mechanism of electron-hole processes of thermoluminescence and thermally stimulated exoelectron emission peaks. Single crystals prepared in highly reducing conditions were treated in different atmospheres (reducing/oxidising) and their optical absorption (after isochronous annealing), fluorescence, positron-annihilation and TL properties were investigated. Changes of optical absorption and positron annihilation spectra of oxidised samples suggest that Schottky-type defect formation occurred - the concentration of Al-vacancies increased while that of O-vacancies decreased. Also we observed the oxidation of Ti 3+ to Ti 4+ . The reduction of an oxidised sample led to the diminution of Al-vacancy concentration while that of O-vacancies has not been changed. This suggest a Frenkel-type cationic defect. We also tried to correlate these changes to changes of TL glow curves and to explain the origin of some TL peaks

  4. Type testing of the Siemens Plessey electronic personal dosemeter.

    Science.gov (United States)

    Hirning, C R; Yuen, P S

    1995-07-01

    This paper presents the results of a laboratory assessment of the performance of a new type of personal dosimeter, the Electronic Personal Dosemeter made by Siemens Plessey Controls Limited. Twenty pre-production dosimeters and a reader were purchased by Ontario Hydro for the assessment. Tests were performed on radiological performance, including reproducibility, accuracy, linearity, detection threshold, energy response, angular response, neutron response, and response time. There were also tests on the effects of a variety of environmental factors, such as temperature, humidity, pulsed magnetic and electric fields, low- and high-frequency electromagnetic fields, light exposure, drop impact, vibration, and splashing. Other characteristics that were tested were alarm volume, clip force, and battery life. The test results were compared with the relevant requirements of three standards: an Ontario Hydro standard for personal alarming dosimeters, an International Electrotechnical Commission draft standard for direct reading personal dose monitors, and an International Electrotechnical Commission standard for thermoluminescence dosimetry systems for personal monitoring. In general, the performance of the Electronic Personal Dosemeter was found to be quite acceptable: it met most of the relevant requirements of the three standards. However, the following deficiencies were found: slow response time; sensitivity to high-frequency electromagnetic fields; poor resistance to dropping; and an alarm that was not loud enough. In addition, the response of the electronic personal dosimeter to low-energy beta rays may be too low for some applications. Problems were experienced with the reliability of operation of the pre-production dosimeters used in these tests.

  5. Principle and use of dosemeters

    International Nuclear Information System (INIS)

    Genet, S.; Ribaud, I.

    1997-01-01

    The institute of nuclear physics of Orsay is accepted by the Ministry of Labour to insure the surveillance of workers individual external exposure against the danger of ionizing radiations and to execute monitoring of radiation protection. The dosemeters studied in this report are the photographic film dosemeters ( for x rays, gamma rays and beta - rays), the trace detectors (fast neutrons) and the radio thermo luminescent dosemeters. For each type, the principle and how to use it are given. (N.C.)

  6. Thermoluminescence dosimetry materials: properties and uses

    International Nuclear Information System (INIS)

    McKeever, S.W.S.; Moscovitch, M.; Townsend, P.D.

    1995-01-01

    This book selects a range of the most popular thermoluminescence dosemeter (TLD) materials in use today and provides a critical account of their thermoluminescence (TL) and dosimetric properties. The information provided includes in-depth discussions of TL mechanisms, including an account of luminescence properties, and relevant information regarding dosimetric characteristics. The book is intended for those involved in TLD materials research, and for technicians and workers involved in the practical application of these materials in TL dosimetry. The advent of modern spectroscopic methods for measuring TL emission spectra (the so-called ''3-D'' presentation) seemed to the authors to be an invitation to compile such spectra for all the major TLD materials. Further consideration led to an expansion of the initial idea to include a compilation of dosimetric properties. One intention is to provide a synopsis of the TL and dosimetric properties of the most widely used TLD materials currently available and to form a link between the solid state defect properties of these materials and their actual dosimetric properties. A second intention is to provide a solid framework from which future studies of TLD materials could be launched. Too often in the past research into TLD materials has been haphazard, to say the least. By illustrating the links between solid state physics and the radiation dosimetry properties of these materials the book points to the future and to the pressing need for enhanced research on TLD materials. (Author)

  7. Thermoluminescence dosimetry properties and kinetic parameters of lithium potassium borate glass co-doped with titanium and magnesium oxides

    International Nuclear Information System (INIS)

    Hashim, S.; Alajerami, Y.S.M.; Ramli, A.T.; Ghoshal, S.K.; Saleh, M.A.; Abdul Kadir, A.B.; Saripan, M.I.; Alzimami, K.; Bradley, D.A.; Mhareb, M.H.A.

    2014-01-01

    Lithium potassium borate (LKB) glasses co-doped with TiO 2 and MgO were prepared using the melt quenching technique. The glasses were cut into transparent chips and exposed to gamma rays of 60 Co to study their thermoluminescence (TL) properties. The TL glow curve of the Ti-doped material featured a single prominent peak at 230 °C. Additional incorporation of MgO as a co-activator enhanced the TL intensity threefold. LKB:Ti,Mg is a low-Z material (Z eff =8.89) with slow signal fading. Its radiation sensitivity is 12 times lower that the sensitivity of TLD-100. The dose response is linear at doses up to 10 3 Gy. The trap parameters, such as the kinetics order, activation energy, and frequency factor, which are related to the glow peak, were determined using TolAnal software. - Highlights: • Lithium potassium borate glass doped with Ti and Mg was prepared. • The material is close to soft tissues in terms of Zeff. • The radiation sensitivity is about 12 times lower than that of TLD-100. • The signal fades about 8% in 10 days and 17% in 3 months

  8. Microcomputer-controlled thermoluminescent analyser IJS MR-200

    International Nuclear Information System (INIS)

    Mihelic, M.; Miklavzic, U.; Rupnik, Z.; Satalic, P.; Spreizer, F.; Zerovnik, I.

    1985-01-01

    Performances and concept of the multipurpose, microcomputer-controlled thermoluminescent analyser, designed for use in laboratory work TL dosemeters as well as for routine dose readings in the range from ecological to accident doses is described. The main features of the analyser are: time-linear sampling, digitalisation, storing, and subsequent displaying on the monitor time scale of the glow and and temperature curve of the TL material; digital stabilization, control and diagnostic of the analog unit; ability of storing 7 different 8-parametric heating programs; ability of storing 15 evaluation programs defined by 2 or 4 parameters and 3 different algorithms (altogether 5 types of evaluations). Analyser has several features intended for routine work: 9 function keys and possibilities of file forming on cassette or display disc, of dose calculation and averaging, of printing reports with names, and possibility of additional programming in Basic. (author)

  9. Response of electret dosemeter to slow neutrons

    International Nuclear Information System (INIS)

    Ghilardi, A.J.P.; Pela, C.A.; Zimmerman, R.L.

    1987-01-01

    The response of electret dosemeter to slow neutrons exposure is cited, mentioning the preparation and the irradiation of dosemeter with Am-Be source. Some theory considerations about the response of electret dosemeter to slow and fast neutrons are also presented. (C.G.C.) [pt

  10. External dosimetry by Thermoluminescent Dosimetry Laboratory - IPEN/CNEN - Brazil

    International Nuclear Information System (INIS)

    Manzoli, Jose Eduardo; Carvalho, Ricardo Nunes

    2001-01-01

    The Thermoluminescence Dosimetry Laboratory, LDT - IPEN/CNEN - Brazil, executes around six thousand TLD readings monthly. It is responsible for the dose evaluation and registration and for all the manufacture process of preparation and sending the dosemeters which are used by IPEN itself, hospitals, industries, town halls, universities, etc. Although prepared to read many kinds of materials, actually the LDT reads only CaSO 4 :Dy detector chips. Chips, heating molds, plastic packaging and almost every material or instrument are made right here at IPEN. The readings are performed in a Bicron Harshaw TLD model 5500, but an old model 2000-AB is ready to be used if necessary. In this work it is presented some features of this important service realized to the IPEN workers and to the community. (author)

  11. EDEL: ENEA dosemeter for eye lens

    International Nuclear Information System (INIS)

    Ferrari, Paolo; Mariotti, Francesca; Campani, Lorenzo

    2016-01-01

    Since the publication of International Commission on Radiological Protection statement in 2011 on tissue reaction, eye lens radiation protection played an important role in exposed personnel dosimetry. For this reason, the Italian National Agency for New Technologies, Energy and Sustainable Economic Development (ENEA) Individual Monitoring Service decided to study a prototype to fulfil specific requests (e.g. for survey in interventional department and intercomparisons). On the basis of such preliminary investigation, a new eye lens dosemeter was developed. The new dosemeter, named EDEL (ENEA Dosemeter for Eye Lens), was characterised in terms of H p (3), the operational quantity related to eye lens monitoring. The investigation was performed experimentally and optimised using the Monte Carlo MCNP6 code. The new prototype was thought to fulfil two main requests: the reliability of the dosimetric data and the portability of the dosemeter itself. The new dosemeter will soon be supplied to the collaborating hospitals for workplace test measurements. (authors)

  12. Development of quality assurance procedures for thermoluminescent CaSO sub 4 :Dy-teflon skived tape dosemeter

    Energy Technology Data Exchange (ETDEWEB)

    Nagpal, J.S.; Popli, K.L.; Kher, R.K.; Varadharajan, G. (Bhabha Atomic Research Centre, Bombay (India). Div. of Radiological Protection)

    1992-01-01

    CaSO{sub 4}:Dy-Teflon TLD dosemeter material has been manufactured locally in a 0.4 mm thick continuous tape form. Fourteen batches have been examined for physical thickness, density, spectral transmission, diffuse optical density, beta transmission and TL output per unit area for evolving a procedure for quality assurance of the TLD element for large scale production. With the exception of two batches, the TL output per unit area of the tape for a number of batches lies within {+-}7% of that for a standard batch. (author).

  13. Technical specification of the NRPB Nuclear Emulsion Dosemeter

    International Nuclear Information System (INIS)

    Bartlett, D.T.; Bird, T.V.

    1978-08-01

    This document is a formal specification of the NRPB Nuclear Emulsion Dosemeter. The dosemeter specified in this report replaces the NRPB Fast Neutron Personal Dosemeter specified in NRPB-R50. (author)

  14. A new extremity dosemeter for beta and gamma radiation

    International Nuclear Information System (INIS)

    Heinzelmann, M.; Pagenkamper, M.

    1988-01-01

    An extremity dosemeter developed at the Juelich Nuclear Research Centre is very well suited for the precise and energy-independent measurement of the skin dose generated by beta or gamma radiation. This is also confirmed by the results of this intercomparison programme. The dosemeter contains three TLDs of LiF in Teflon mounted behind a window of 0.9 mg/cm 2 . The great advantage of this dosemeter is three TLD's enabling statements about the radiation quality. However, the dosemeter has two disadvantages The dosemeter is complicated to manufacture. A very thin plastic foil of 0.9 mg/cm 2 must be attached to a support. This work is difficult and time-consuming and cannot be automated. The window in front of the TLD is not sturdy enough and is occasionally destroyed when the dosemeter is being worn. These two disadvantages prevent this extremity dosemeter from being used more frequently. For this reason, work was begun on developing a new extremity dosemeter without these two disadvantages. The great advantage of the previous dosemeter of obtaining statements about the type of radiation with the aid of readings from three TLD's was to be retained. The improved extremity dosemeter has a more sturdy and thicker window with a similar response as the previous dosemeter with a thinner window

  15. Practical use of LiF-Nuclemon as a thermoluminescent dosemeter of γ, χ and β radiations

    International Nuclear Information System (INIS)

    Marilac Pereira Dolabela, L. de.

    1987-01-01

    The practical use of doped LiF Nuclemon with magnesium (LiF:Mg) as a thermoluminescent dosimeter of γ, X and β radiations is studied. The TL properties of LiF:Mg were analysed to γ energy of 1,25 MeV; 30, 50 and 70 KeV. (author)

  16. Evaluation of the kinetic parameters of CaF2:Tm (TLD-300) thermoluminescence dosemeters

    International Nuclear Information System (INIS)

    Azorin, Juan; Gutierrez, Alicia; Furetta, Claudio

    1989-01-01

    The kinetic parameters of the first, third and fifth peaks of CaF 2 :Tm (TLD-300) dosemeters were determined experimentally using methods based on initial rise, peak shape, different heating rates, isothermal decay and numerical fitting method and a comparison was made between the results obtained. The average activation energy values determined by the methods used here were 0.72, 1.16 and 1.77 eV, with mean values for the frequency factor of 2.58 x 10 8 , 2.89 x 10 13 and 6.16 x 10 16 s -1 for the first, third and fifth peaks respectively. The values of activation energy obtained for each peak using the different methods covered a spread of about 4%. However, the spread for each of the frequency factors was very high. (author)

  17. High-yield production of hydrophobins RodA and RodB from Aspergillus fumigatus in Pichia pastoris

    DEFF Research Database (Denmark)

    Pedersen, Mona Højgaard; Borodina, Irina; Moresco, Jacob Lange

    2011-01-01

    A as well as rRodB were able to convert a glass surface from hydrophilic to hydrophobic similar to native RodA, but only rRodB was able to decrease the hydrophobicity of a Teflon-like surface to the same extent as native RodA, while rRodA showed this ability to a lesser extent. Recombinant RodA and native...

  18. Thermoluminescent dosimetry of beta radiations of 90 Sr/ 90 Y using amorphous ZrO2

    International Nuclear Information System (INIS)

    Rivera M, T.; Olvera T, L.; Azorin N, J.; Barrera R, M.; Soto E, A.M.

    2005-01-01

    In this work the results of studying the thermoluminescent properties (Tl) of the zirconium oxide in its amorphous state (ZrO 2 -a) before beta radiations of 90 Sr/ 90 Y are presented. The amorphous powders of the zirconium oxide were synthesized by means of the sol-gel technique. The sol-gel process using alkoxides like precursors, is an efficient method to prepare a matrix of zirconium oxide by hydrolysis - condensation of the precursor to form chains of Zr-H 3 and Zr-O 2 . One of the advantages of this technique is the obtention of gels at low temperatures with very high purity and homogeneity. The powders were characterized by means of thermal analysis and by X-ray diffraction. The powders of ZrO 2 -a, previously irradiated with beta particles of 90 Sr/ 90 Y, presented a thermoluminescent curve with two peaks at 150 and 257 C. The dissipation of the information of the one ZrO 2 -a was of 40% the first 2 hours remaining constant the information for the following 30 days. The reproducibility of the information was of ± 2.5% in standard deviation. The studied characteristics allow to propose to the amorphous zirconium oxide as thermoluminescent dosemeter for the detection of beta radiation. (Author)

  19. Thermoluminescent dosimetry of beta radiations of 90 Sr/ 90 Y using ZrO2: Eu

    International Nuclear Information System (INIS)

    Olvera T, L.; Azorin N, J.; Barrera S, M.; Soto E, A.M.; Rivera M, T.

    2005-01-01

    In this work the results of studying the thermoluminescent properties (TL) of the doped zirconium oxide with europium (ZrO 2 : Eu 3+ ) before beta radiations of 90 Sr/ 90 Y are presented. The powders of ZrO 2 : Eu 3+ were obtained by means of the sol-gel technique and they were characterized by means of thermal analysis and by X-ray diffraction. The powders of ZrO 2 : Eu 3+ , previously irradiated with beta particles of 90 Sr/ 90 Y, presented a thermoluminescent curve with two peaks at 204 and 292 C respectively. The TL response of the ZrO 2 : Eu 3+ as function of the absorbed dose was lineal from 2 Gy up to 90 Gy. The fading of the information of the ZrO 2 : Eu 3+ was of 10% the first 2 hours remaining almost constant the information by the following 30 days. The ZrO 2 doped with the (Eu 3+ ) ion it was found more sensitive to the beta radiation that the one of zirconium oxide without doping (ZrO 2 ) obtained by the same method. Those studied characteristics allow to propose to the doped zirconium oxide with europium like thermoluminescent dosemeter for the detection of the beta radiation. (Author)

  20. Calibrations of pocket dosemeters using a comparison method

    International Nuclear Information System (INIS)

    Somarriba V, I.

    1996-01-01

    This monograph is dedicated mainly to the calibration of pocket dosemeters. Various types of radiation sources used in hospitals and different radiation detectors with emphasis on ionization chambers are briefly presented. Calibration methods based on the use of a reference dosemeter were developed to calibrate all pocket dosemeters existing at the Radiation Physics and Metrology Laboratory. Some of these dosemeters were used in personnel dosimetry at hospitals. Moreover, a study was realized about factors that affect the measurements with pocket dosemeters in the long term, such as discharges due to cosmic radiation. A DBASE IV program was developed to store the information included in the hospital's registry

  1. Structural Glass Beams with Embedded Glass Fibre Reinforcement

    NARCIS (Netherlands)

    Louter, P.C.; Leung, Calvin; Kolstein, M.H.; Vambersky, J.N.J.A.; Bos, Freek; Louter, Pieter Christiaan; Veer, Fred

    2010-01-01

    This paper investigates the possibilities of pultruded glass fibre rods as embedded reinforcement in SentryGlas (SG) laminated glass beams. To do so, a series of pullout tests, to investigate the bond strength of the rods to the laminate, and a series of beam tests, to investigate the post-breakage

  2. Electret dosemeter for beta radiation

    International Nuclear Information System (INIS)

    Campos, L.L.; Caldas, L.V.E.; Mascarenhas, S.

    The response characteristics of an electret dosemeter for beta radiation are studied. Experiments were performed using different geometries and walls, and it was verified for which geometry the dosemeter sensitivity is greater. Sources of 90 Sr - 90 Y, 204 Tl and 85 Kr were used in the experiments. (I.C.R.) [pt

  3. Comparing different error-conditions in film dosemeter evaluation

    International Nuclear Information System (INIS)

    Roed, H.; Figel, M.

    2007-01-01

    In the evaluation of a film used as a personal dosemeter it may be necessary to mark the dosemeters when possible error-conditions are recognised, such as errors that have an influence on the ability to make a correct evaluation of the dose value. In this project a comparison has been carried out to examine how two individual monitoring services, IMS [National Inst. of Radiation Hygiene (Denmark) (NIRH) and National Research Centre for Environment and Health (Germany) (GSF)], from two different EU countries mark their dosemeters. The IMS are different in size, type of customers and issuing period, but both use films as their primary dosemeters. The error-conditions examined are dosemeters exposed to moisture or light, contaminated dosemeters, films exposed outside the badge, missing filters in the badge, films inserted incorrectly in the badge and dosemeters not returned or returned too late to the IMS. The data are collected for the year 2003 where NIRH evaluated ∼50,000 and GSF ∼1.4 million film dosemeters. The percentage of film dosemeters is calculated for each error-condition as well as the distribution among eight different employee categories, i.e. medicine, nuclear medicine, nuclear industry, industry, radiography, laboratories, veterinary and others. It turned out, that incorrect insertion of the film in the badge was the most common error-condition observed at both IMS and that veterinarians, as the employee category, generally have the highest number of errors. NIRH has a significantly higher relative number of dosemeters in most error-conditions than GSF, which perhaps reflects that a comparison is difficult due to different systemic and methodical differences between the IMS and countries, e.g. regulations and monitoring programs etc. Also the non-existence of a common categorisation method for employee categories contributes to make a comparison like this difficult. (authors)

  4. In vivo thermoluminescent dosimetry in studies of helicoid computed tomography and excretory urogram

    International Nuclear Information System (INIS)

    Cruz C, D.; Azorin N, J.; Saucedo A, V.M.; Barajas O, J.L.

    2005-01-01

    The dosimetry is the field of measurement of the ionizing radiations. It final objective is to determine the 'absorbed dose' for people. The dosimetry is vital in the radiotherapy, the radiological protection and the treatment technologies by irradiation. Presently work, we develop 'In vivo' dosimetry, in exposed patients to studies of helical computed tomography and excretory urogram. The dosimetry 'in vivo' was carried out in 20 patients selected aleatorily, for each medical study. The absorbed dose was measured in points of interest located in crystalline, thyroid, chest and abdomen of each patient, by means of thermoluminescent dosemeters (TLD) LiF: Mg,Cu,P + Ptfe of national fabrication. Also it was quantified the dose in the working area. (Author)

  5. Czech results at criticality dosimetry intercomparison 2002.

    Science.gov (United States)

    Frantisek, Spurný; Jaroslav, Trousil

    2004-01-01

    Two criticality dosimetry systems were tested by Czech participants during the intercomparison held in Valduc, France, June 2002. The first consisted of the thermoluminescent detectors (TLDs) (Al-P glasses) and Si-diodes as passive neutron dosemeters. Second, it was studied to what extent the individual dosemeters used in the Czech routine personal dosimetry service can give a reliable estimation of criticality accident exposure. It was found that the first system furnishes quite reliable estimation of accidental doses. For routine individual dosimetry system, no important problems were encountered in the case of photon dosemeters (TLDs, film badge). For etched track detectors in contact with the 232Th or 235U-Al alloy, the track density saturation for the spark counting method limits the upper dose at approximately 1 Gy for neutrons with the energy >1 MeV.

  6. Development of an algorithm for X-ray exposures using the Panasonic UD-802A thermoluminescent dosemeter

    International Nuclear Information System (INIS)

    McKittrick, Leo; Currivan, Lorraine; Pollard, David; Nicholls, Colyn; Romero, A.M.; Palethorpe, Jeffrey

    2008-01-01

    Full text: As part of its continuous quality improvement the Dosimetry Service of the Radiological Protection Institute of Ireland (RPII) in conjunction with Panasonic Industrial Europe (UK) has investigated further the use of the standard Panasonic algorithm for X-ray exposures using the Panasonic UD-802A TL dosemeter. Originally developed to satisfy the obsolete standard ANSI 13.11-1983, the standard Panasonic dose algorithm has undergone several revisions such as HPS N13.11-2001. This paper presents a dose algorithm that can be used to correct the dose response at low energies such as X-ray radiation using a four element TL dosemeter due to the behaviour of two different independent phosphors. A series of irradiations with a range of energies using N-20 up to Co-60 were carried out with our particular interest being in responses to X-ray irradiations. Irradiations were performed at: RRPPS, University Hospital Birmingham NHS Foundation Trust, U.K.; HPA, U.K. and CIEMAT, Madrid, Spain. Different irradiation conditions were employed which included: X-ray from narrow and wide spectra as described by ISO 4037-1 (1996), and ISO water slab phantom and PMMA slab phantom respectively. Using the UD-802A TLD and UD-854AT hanger combination, the response data from the series of irradiations was utilised to validate and if necessary, modify the photon/beta branches of the algorithm to: 1. Best estimate Hp(10) and Hp(0.07); 2. Provide information on irradiation energies; 3. Verification by performance tests. This work further advances the algorithm developed at CIEMAT whereby a best-fit, polynomial trend is used with the dose response variations between the independent phosphors. (author)

  7. Radiation exposure of an anaesthesiologist in catheterisation and electrophysiological cardiac procedures

    International Nuclear Information System (INIS)

    Andreoli, Stefano; Moretti, Renzo; Lorini, Ferdinando Luca; Lagrotta, Mariavittoria

    2016-01-01

    Sometimes, cardiac catheterisation and electrophysiological procedures, diagnostic and interventional, require an anaesthesiological support. The anaesthesiologist receives radiation doses depending on various factors, such as type of procedure and exposure modality, anaesthesiological technique, individual protective devices and operator experience. The aim of this study was to investigate the dose per procedure, the exposure inhomogeneity and the effective dose, E, of a senior anaesthesiologist in the haemodynamic laboratory of Ospedali Riuniti, Bergamo. The dose monitoring was routinely performed with sets of several thermoluminescent dosemeters and an electronic personal dosemeter. The study covered 300 consecutive procedures over 1 y. The anaesthesiologist wore a protective apron, a thyroid collar and glasses (0.5 mm lead-equivalent). (authors)

  8. Thermoluminescence and optically stimulated luminescence properties of Dy3+-doped CaO-Al2O3-B2O3-based glasses

    Science.gov (United States)

    Yahaba, T.; Fujimoto, Y.; Yanagida, T.; Koshimizu, M.; Tanaka, H.; Saeki, K.; Asai, K.

    2017-02-01

    We developed Dy3+-doped CaO-Al2O3-B2O3 based glasses with Dy concentrations of 0.5, 1.0, and 2.0 mol% using a melt-quenching technique. The as-synthesized glasses were applicable as materials exhibiting thermoluminescence (TL) and optically stimulated luminescence (OSL). The optical and radiation response properties of the glasses were characterized. In the photoluminescence (PL) spectra, two emission bands due to the 4F9/2 → 6H15/2 and 4F9/2 → 6H13/2 transitions of Dy3+ were observed at 480 and 580 nm. In the OSL spectra, the emission band due to the 4F9/2 → 6H15/2 transition of Dy3+ was observed. Excellent TL and OSL responses were observed for dose ranges of 0.1-90 Gy. In addition, TL fading behavior was better than that of OSL in term of the long-time storage. These results indicate that the Dy3+-doped CaO-Al2O3-B2O3-based glasses are applicable as TL materials.

  9. Passive neutron dosemeter with activation detector

    Energy Technology Data Exchange (ETDEWEB)

    Valero L, C.; Banuelos F, A.; Guzman G, K. A.; Borja H, C. G.; Hernandez D, V. M.; Vega C, H. R., E-mail: fermineutron@yahoo.com [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Calle Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico)

    2011-10-15

    A passive neutron dosemeter with {sup 197}Au activation detector has been developed. The area dosemeter was made as a 20.5 {phi} x 20.5 cm{sup 2} polyethylene moderator, with a polyethylene pug where a {sup 197}Au foil can be located either parallel or perpendicular to moderator axis. Using Monte Carlo methods, with the MCNP5 code. With the fluence response and the fluence-to-equivalent dose conversion coefficients from ICRP-74, responses to H*(10) were also calculated, these were compared against responses of commercially available neutron area monitors and dosemeters. (Author)

  10. Laser amplifier based on a neodymium glass rod 150 mm in diameter

    Energy Technology Data Exchange (ETDEWEB)

    Shaykin, A A; Fokin, A P; Soloviev, A A; Kuzmin, A A; Shaikin, I A; Burdonov, K F; Khazanov, E A [Institute of Applied Physics, Russian Academy of Sciences, Nizhnii Novgorod (Russian Federation); Charukhchev, A V [Public Limited Company " Scientific research Institute for Optoelectronic Instrument Engineering" , Leningrad region (Russian Federation)

    2014-05-30

    A unique large-aperture neodymium glass rod amplifier is experimentally studied. The small-signal gain distribution is measured at different pump energies. The aperture-averaged gain is found to be 2.3. The stored energy (500 J), the maximum possible pump pulse repetition rate, and the depolarisation in a single pulse and in a series of pulses with a repetition rate of one pulse per five minutes are calculated based on the investigations performed. It is shown that the use of this amplifier at the exit of the existing laser can increase the output pulse energy from 300 to 600 J. (lasers)

  11. Evaluation of the glow curves of a new glass matrix

    International Nuclear Information System (INIS)

    Oliveira, Nathália S.; Souza, Samara P.; Ferreira, Pâmela Z.; Dantas, Noelio O.; Silva, Anielle C.A.; Neves, Lucio P.; Perini, Ana P.; Carrera, Betzabel N.S.; Watanabe, Shigueo

    2017-01-01

    Thermoluminescence is a dosimetric technique with may be used to personal, clinical, environmental and high doses. In this work a new glass matrix, with nominal composition of 20Li 2 CO 3 .10Al 2 O 3 .25BaO.45B 2 O 3 (mol%), was studied by the thermoluminescence technique. The glow curves was be analyzed, after the irradiation of this glass matrix with high doses. The results showed that this new glass matrix has a temperature peak in 260°C, which is ideal for dosimetry applications. (author)

  12. Mensuration of equivalent dose with personal dosemeters and instruments of radiological protection in the new operative magnitudes ICRU, for external fields of beta radiation. Part I. Study of the homogeneity of the response personal dosemeters leaves (cards G-1, TLD-100), in radiation fields of Co60

    International Nuclear Information System (INIS)

    Alvarez R, J.T.

    1994-01-01

    without personal dosemeters, and for the bottom readings and sensibility of the reader equipment; and of the adjustment kindness to the normal for each game of data.The main conclusions that they were derived of this study they are: i.That the reader equipment presents differences statistically significant (α=5%) in their sensibilities; that which is necessary correct the readings of the dosemeter, making stand out the necessity of building the curves of operation of the equipment for the bottom and the sensibility. ii. The filter of A1 is not equivalent among the different personal dosemeters i i.For this sample of 40 cards with personal dosemeters to the one fewer exist: eleven populations different statistic significant, (α) to the response glass 1 five for the response of the glass 2, respectively. i v. That exist difference significant statistical, (α =5%) among the repetitions; what implies that it is necessary to control but strictly the experimental parameters to the one moment of carrying out the irradiations, (field size, correction for time of obturator of the source, distances source-detecting, etc), since the influence of the reader equipment it has already been considered. (Author)

  13. Rethinking About Chronology Of Chichen Itza: By Thermoluminescence Dating Of Volcanic Glass.

    Science.gov (United States)

    Chung, H.; Gonzalez, P.; Ramirez, A.; Schaaf, P.; Lee, I.

    One of the most frequently recurring questions in archaeology concerns the age of the studied objects. To date, determining the chronology of the Puuc and Chichen Slate pottery of Chchen Itza, based on the stylistic change in the pottery, has been problematic. The Thermoluminescence method of dating pottery was applied to samples of Chichen Itza, Yucatan, Mexico, which contain volcanic glass as temper. They were analyzed using the fine grain technique. The radisotopes that contribute to the accumulate annual dose in ceramic samples (40K, 238U, 232Th) were determined by means of Energy Dispersive X-ray Spectroscopy and Neutron Activation Analysis technique, while the artificial irradiation of the samples was carried out using 90Sr source beta radiation. We obtained results indicating the following dates; 875 ± 88; 1055 ± 85; 1063 ± 47, 1154 ± 76; 1110 ± 53, 1132 ± 69, 1221 ± 30, 1532 ± 26. The results demonstrate a chronological order of dates. Most results are in accordance with the established dates of Chichen Itza. However, Puuc and Chichen Slate did not show differences of 200 years as previously proposed, and there is evidence of an earlier period than those two pottery periods. This necessitates a reconsideration of the chronology of Chichen Itza.

  14. The need for comparison of diagnostic dosemeters

    International Nuclear Information System (INIS)

    Kramer, H.M.

    1989-01-01

    One precondition for successfully implementing quality control in diagnostic radiology is that the measuring instruments used are of adequate quality. Among the measuring instruments used for quality control, dosemeters play a predominant role. Therefore it is essential to know the performance characteristics of a suitable dosemeter and to have information on the uncertainties and measurement errors found during acceptance testing. An outline will be given of the requirements on the accuracy of dosemeters. Such requirements have to address the task of the dosemeter and also take into account their technical feasibility. In order to gather information on the performance characteristics of field instruments, a programme for a European intercomparison has been developed which was carried out in 1988. The programme, together with its underlying ideas and its objectives, is presented in this paper. (author)

  15. On optically stimulated luminescence properties of household salt as a retrospective dosemeter

    International Nuclear Information System (INIS)

    Timar-Gabor, A.; Trandafir, O.

    2013-01-01

    Thermoluminescence (TL) and optically stimulated luminescence (OSL) in the UV (270-370 nm) spectral region have been investigated for five types of table salt (NaCl) available in Romanian supermarkets with a view to applying them in retrospective dosimetry. The salt samples gave bright TL signals with two main peaks at ∼ 100 deg. C and at 300 or 260 deg. C, depending on the origin of the salt and bright OSL signals under continuous stimulation with blue light. The OSL signal (stimulated at 100 deg. C after a pre-heat of 10 s at 150 deg. C) was used for investigations in a standard multiple aliquot procedure. The dose- response was found to be linear in the dose range investigated (up to ∼ 100 mGy) and the lower limit of detection for the samples varied from ∼ 0.01 to 14 mGy. These characteristics, along with the widespread abundance and low cost of household salt, confirm its potential as a retrospective dosemeter. (authors)

  16. New ISO standard - personnel photographic film dosemeters

    International Nuclear Information System (INIS)

    Brabec, D.

    1980-01-01

    The ISO Standard 1757 ''Personnel Photographic Film Dosemeters'', issued in June 1980, is briefly described. UVVVR's own dosemeter developed for use in the national film dosimetry service in Czechoslovakia is evaluated in relation to this ISO Standard. (author)

  17. Feasibility of a neutron detector-dosemeter based on single-event upsets in dynamic random-access memories

    International Nuclear Information System (INIS)

    Phillips, G.W.; August, R.A.; Campbell, A.B.; Nelson, M.E.; Guardala, N.A.; Price, J.L.; Moscovitch, M.

    2002-01-01

    The feasibility was investigated of a solid-state neutron detector/dosemeter based on single-event upset (SEU) effects in dynamic random-access memories (DRAMs), commonly used in computer memories. Such a device, which uses a neutron converter material to produce a charged particle capable of causing an upset, would be light-weight, low-power, and could be read simply by polling the memory for bit flips. It would have significant advantages over standard solid-state neutron dosemeters which require off-line processing for track etching and analysis. Previous efforts at developing an SEU neutron detector/dosemeter have suffered from poor response, which can be greatly enhanced by selecting a modern high-density DRAM chip for SEU sensitivity and by using a thin 10 B film as a converter. Past attempts to use 10 B were not successful because the average alpha particle energy was insufficient to penetrate to the sensitive region of the memory. This can be overcome by removing the surface passivation layer before depositing the 10 B film or by implanting 10B directly into the chip. Previous experimental data show a 10 3 increase in neutron sensitivity by chips containing borosilicate glass, which could be used in an SEU detector. The results are presented of simulations showing that the absolute efficiency of an SEU neutron dosemeter can be increased by at least a factor of 1000 over earlier designs. (author)

  18. Fuel rod simulator effects in flooding experiments single rod tests

    International Nuclear Information System (INIS)

    Nishida, M.

    1984-09-01

    The influence of a gas filled gap between cladding and pellet on the quenching behavior of a PWR fuel rod during the reflood phase of a LOCA has been investigated. Flooding experiments were conducted with a short length electrically heated single fuel rod simulator surrounded by glass housing. The gap of 0.05 mm width between the Zircaloy cladding and the internal Al 2 O 3 pellets of the rod was filled either wit helium or with argon to vary the radial heat resistance across the gap. This report presents some typical data and an evaluation of the reflood behavior of the fuel rod simulator used. The results show that the quench front propagates faster for increasing heat resistance in the gap between cladding and heat source of the rod. (orig.) [de

  19. Response of electret dosemeter to slow neutrons

    International Nuclear Information System (INIS)

    Ghilardi, A.J.P.; Pela, C.A.; Zimmerman, R.L.

    1987-01-01

    The response of the electret dosemeter to exposition of slow neutrons is studied. Different external coatings are used on the dosemeter (polyethylene, alminium, polyethylene + boron, aluminium + boron) and exposure curves (with and without water) are compared. (M.A.C.) [pt

  20. Preliminary study on rotary ultrasonic machining of Bk-7 optical glass rod

    International Nuclear Information System (INIS)

    Hamzah, E.; Izman, S.; Khoo, C.Y.; Zainal Abidin, N.N.

    2007-01-01

    This paper presents an experimental observation on rotary ultrasonic machining (RUM) of BK7 optical glass rod. BK7 is a common technical optical glass for high quality optical components due to its high linear optical transmission in the visible range and is chemically stable. RUM is a hybrid machining process that combines the material removal mechanisms of diamond grinding and ultrasonic machining (USM) and it is non-thermal, non-chemical, creates no change in the microstructure, chemical or physical properties of the work piece. In the RUM, a controlled static load is applied to the rotating core drill with metal bonded diamond abrasive and is ultrasonically vibrated in the axial direction. A water-soluble coolant was used to cool the tool and sample during machining processes. By using DOE (Design of Experiment) approach, the effect of spindle speed and feed rate to the ultrasonic machinability had been developed. The main effects and two-factor interactions of process parameters (spindle speed) and feed rate) on output variables (MRR, surface roughness, opaqueness, chipping thickness and chipping size) are studied. (author)

  1. Application of spectroscopic techniques in the radiation dosimetry of glasses: An update

    International Nuclear Information System (INIS)

    Natarajan, V

    2009-01-01

    The colorimetry and thermoluminescence properties of gamma irradiated glass were reported in as early as 1920. The utility of radio-photoluminescence (RPL) of silver activated metaphosphate glass for monitoring high doses of accidental and routine gamma radiation was reported in the 1960s. Since then considerable amount of research work has been carried out to study the thermoluminescence (TL), optical absorption (OA), electron paramagnetic resonance (EPR) and optically stimulated luminescence (OSL) of different commercially available glasses for high as well as low dose applications. A brief review of the progress made in the spectroscopic studies of glasses during the past few decades and the application of glasses for radiation dosimetry has been given in this paper.

  2. Application of spectroscopic techniques in the radiation dosimetry of glasses: An update

    Energy Technology Data Exchange (ETDEWEB)

    Natarajan, V, E-mail: vnatra@yahoo.co.in [Radiochemistry Division, Bhabha Atomic Research Centre, Mumbai 40085 (India)

    2009-07-15

    The colorimetry and thermoluminescence properties of gamma irradiated glass were reported in as early as 1920. The utility of radio-photoluminescence (RPL) of silver activated metaphosphate glass for monitoring high doses of accidental and routine gamma radiation was reported in the 1960s. Since then considerable amount of research work has been carried out to study the thermoluminescence (TL), optical absorption (OA), electron paramagnetic resonance (EPR) and optically stimulated luminescence (OSL) of different commercially available glasses for high as well as low dose applications. A brief review of the progress made in the spectroscopic studies of glasses during the past few decades and the application of glasses for radiation dosimetry has been given in this paper.

  3. Colour dosemeters for high level radiation dosimetry

    DEFF Research Database (Denmark)

    Schönbacher, H.; Coninckx, F.; Miller, A.

    1990-01-01

    interpretation or a dose measurement with a simple instrument such as a portable reflecting densitometer in the range of 10(3) to 10(6) Gy. Two projects were investigated: (1) a thin plastic film with a self adhesive tape containing a radiochromic dye which induces a colour change when exposed to ionising...... of paint and film dosemeters were installed in the 450 GeV Super Proton Synchrotron (SPS) at CERN and irradiated during operation for more than two years. Within the useful range of the dosemeters, dose comparisons with other dosemeter types gave satisfactory results. Application in other fields...

  4. Evaluation of BICRON NE MCP DXT-RAD passive extremity dosemeter

    CERN Document Server

    Yuen, P S; Frketich, G; Rotunda, J

    1999-01-01

    Passive extremity dosemeters currently used in dosimetry communities worldwide have shortcomings. In general, an extremity dosemeter has too thick a detector element, and the dosemeter response is highly energy dependent for beta rays with energies ranging from 200 keV to 2 MeV. It often does not have dosemeter identification, causing problems in the chain of custody. It is often read manually, rendering reading/packing operations very labour intensive. As a result of collaboration between AECL and BICRON NE, a new extremity dosemeter, incorporating a highly sensitive LiF:Mg,Cu,P TLD and tentatively code named MCP DXT-RAD, was developed. It has been evaluated for radiological performance against an ISO draft standard for extremity dosemeters in twelve categories: homogeneity, detection threshold, beta ray energy response, beta angular response, photon energy response, photon angular response, reproducibility, stability under various climatic conditions, linearity, residue, self irradiation, and effect of ligh...

  5. Dating of the archaeological site 'El Tigre' by the thermoluminescent method; Fechamiento del sitio arqueologico 'El Tigre' por el metodo de termoluminiscencia

    Energy Technology Data Exchange (ETDEWEB)

    Portilla, R. De la [ENAH, Periferico Sur y Zapote, 14030 Mexico D.F. (Mexico); Gonzalez, P.R.; Mendoza, D. [ININ, 52750 La Marquesa, Ocoyoacac, Estado de Mexico (Mexico); Vargas, E. [Instituto de Investigaciones Antropologicas, UNAM, 04510 Mexico D.F. (Mexico); Ramirez, A. [Instituto de Geofisica, UNAM, 04510 Mexico D.F. (Mexico)

    2007-07-01

    Full text: The dating of pre hispanic pottery, is supported by techniques such as stratigraphy, typology, in physical and chemical procedures, as the analysis of {sup 14}C and the thermoluminescence (TL). The last one permits us obtain absolute dating of archaeological pottery with an acceptable precision. In this work we apply the applied the thermoluminescent technique to verify the age of the Pre-Classic and Classic Terminal periods proposed for the archaeological site 'El Tigre', Campeche. The samples were obtained during a work period and the radiation of the ground of background (emitted by the ground), as well as the contribution of the cosmic radiation was measured with thermoluminescent dosemeters of LiF: Mg,Cu,P + PTFE, put in the sample zone. The preliminary results indicate that not all the analyzed samples can be dated. The viability of the dating of the samples is discussed based on the fact and function of the presence of certain crystalline phases such as calcite and quartz. The discussion is complemented emphsizing the importance of handling the conditions of handling of samples. (Author)

  6. Factors affecting polyamide prototypes design of Albedo dosemeters

    International Nuclear Information System (INIS)

    Martins, M.M.; Mauricio, C.L.P.; Fonseca, E.S.

    1996-01-01

    This work studies the most important factors which affect the response of albedo neutron dosemeters containing LiF TLDs with the aim to improve their sensitivity. It includes tests of thickness and shape of the polyamide moderator body prototypes, albedo window diameter and TLD position inside the moderator. Analyzing the results, an albedo neutron dosemeter prototype, B 4 C covered, was developed. The prototype has a response three times higher than the albedo dosemeter now in use in Brazil. (author)

  7. Thermoluminescent dosemeters for determining the energy absorbed during X-ray radiography of the vertebral column

    International Nuclear Information System (INIS)

    Liebl, R.

    1983-01-01

    The dose and absorbed energy during normal diagnostic X-ray of various sections of the vertebral column were determined with LiF-dosemeters in a phantom. The paper describes a method to be used to determine integral doses from the dose measurements. The energy absorbed for one X-ray picture of the vertebral column is between 5 and 30 mJ. Compared to other diagnostical X-rays the quantity of the energy absorbed during X-ray of the vertebral column is rather high and is only reached by X-rays in the pelvic region. The speculations on the rate of incidence of malignent neoplasms on the basis of diagnostical X-ray of the vertebral column reveal a value of 50 per 60 x 10 6 persons. This value is likely to overestimate the risk, seems, however, to be low in comparison to other risks of every day life (traffic accident, mountainering, etc). (orig./HP) [de

  8. Comments by N.R.P.B

    International Nuclear Information System (INIS)

    Marshall, T.O.

    1984-01-01

    NRPB operates both photographic film and thermoluminescent dosemeter services for individual monitoring of external sources of photon and beta radiations. Both services issue about 300,000 dosemeters per year. The film dosemeter uses the Eastman Kodak Type 2 film in the NRPB/AERE multi-filter holder and is designed to measure the dose equivalent at or near the surface of the body. The thermoluminescent dosemeter uses two LiF in teflon disks of different thicknesses retained in an encoded aluminium plate which is wrapped and labelled and worn in a plastic holder. The photographic film service is partly automated but does not include a dose record keeping service. The dosemeters used for the intercomparison were taken from our normal stock and were evaluated in the normal routine manner

  9. EURADOS intercomparisons on whole body and extremity dosemeters (2008–2009) – Results and comparison of different dosemeter designs

    International Nuclear Information System (INIS)

    Stadtmann, H.; Grimbergen, T.W.M.; Figel, M.; Romero, A.M.; Mcwhan, A.F.

    2011-01-01

    The EURADOS (European Radiation Dosimetry Group) working group (WG2) on Harmonisation of Individual Monitoring in Europe has shown that intercomparisons are fundamental for harmonisation of individual monitoring. As a result of these considerations, EURADOS started to prepare a series of continuous intercomparisons for individual monitoring services (IMS) in Europe. EURADOS has now successfully executed two intercomparison exercises, one for whole body dosemeters (2008) and one for extremity dosemeters (2009). Both exercises were performed without external funding, all costs being covered by the participants’ fees. More than 120 (!) different – mainly passive – dosimetry systems were tested from all over Europe until now and the results were analysed and compared. Although scientific studies were not the primary objective of these exercises specific additional information about the tested systems supplied by the participants for statistical analysis allows a detailed analysis of these results with respect to different parameters like dosemeter type, detector material, filter thickness or others design parameters. The influence of such parameters on the response values of the dosemeters is analysed and discussed in this paper. Both energy response as well as directional response for beta and photon radiation are considered. The influence of different design parameters e.g. detector material, is analysed by comparing the frequency distribution of the response values for different dosemeter groups. Pronounced differences for different detector materials (LiF:Mg,Ti, LiF:Mg,Cu,P and Li 2 B 4 O 7 /CaSO 4 ) were found. The paper gives a comprehensive overview on the performance of various European dosimetry services and the influence of the dosemeter design on the resulting response values.

  10. Thermal behaviour of the thermoluminescent dosemeter CaSO4: Dy + NaCl

    International Nuclear Information System (INIS)

    Antonio Filho, J.

    1988-01-01

    Thermoluminescent materials are widely applied in radiation dosimetry. Nevertheless to obtain a good degree of reliability, it is necessary to subject the materials to an adequate thermal treatment, before and after irradiation, conveniently controlling temperature, heating time and cooling rate. It has been observed that the TLD CaSO 4 : Dy + NaCl loses 10% of its sensivity after each utilization cycle with the thermal treatment usually recomended. This fact establishes a severe limitation in works requiring a good measurement precision. Several alternative procedures, which overcome this limitation, have been studied. Two of the procedures yielded excellent results, with a coefficient of variation of 1 to 1.3%. One of them consist of two heating cycles at 300 0 C during 15 and three hours respectively, before irradiation, followed by a heating cycle at 100 0 C during 15 minutes, after irradiation. In a second procedure the material is heated at 300 0 C during three hours, pre-irradiation, followed by a heating cycle at 100 0 C. during 15 minutes, after irradiation. In all cases the heating cycle are followed by fast cooling periods. (author) [pt

  11. Mensuration of equivalent dose with personal dosemeters and instruments of radiological protection in the new operative quantities ICRU, for external fields of beta radiation. Part II. I study of the angular response of personal dosemeters TLD-100 in secondary patron fields of beta radiation (90Sr / 90Y)

    International Nuclear Information System (INIS)

    Alvarez R, J.T.

    1994-01-01

    The objective of this work is to carry out one of the possible ones test type for personal dosemeters TLD, under the recomendations of the ICRU 39, ICRU 43 and the draft of the norm ISO 6980,(1992), with the purpose of verifying the capacity of these detectors to carry out the operative unit: H' (0.07;α). Since H' (O. 07;α) this defined one in an expanded field, one of these tests type consist on determining the angular response of these detectors. 20 personal dosemeters TLD-100 was used, (card marks: Harshaw, Model: G-1, with two glasses of TLD-100 absorbed in teflon; the portadosemeters has two windows, a free one and another with a filter of Pb of 171.0 mg cm -2 ); these dosemeters they were previously selected, [to see, S tudy of the Homogeneity of the response of Personal Dosemeters (Cards G-l, TLD-100) in Radiation of Countrysides of 60 Co , J.T. Alvarez R. Technician Report GSR/IT/0001/94].The irradiations to effectued in secondary countryside of radiation beta of 90 Sr/ 90 Y. The study was undertaken by means of an experimental design of blocks random that contemplate the following variables: intensity of the radiation source, (1850 MBq and 74 MBq); position of irradiation, (four positions); incidence of angle of the radiation (0, 15, 30, 45, 60 and 75 grades) and the absorbed dose in air, (0.005, 0.010, 0.020, 0.050 and 0.100 Gy). Then null hypothesis it was to suppose that there was not difference among the stockings of each treatment, to used the statistical of Duncan to carry out tests of stockings at a level of significance of 5%.These tests of stockings throw the following results in those variables of the experimental design: The irradiations carried out so much with the source pattern secondary of 90 Sr/ 90 Y of 1850 MBq and of 74 MBq, they are equivalent reason why they can be used indistinctly. The responses of each one of the glasses of the card are strongly anisotropic for each glass; four positions of irradiation is used: glass 1 (window of

  12. Influence of different types of phantoms on the calibration of dosemeters for eye lens dosimetry

    International Nuclear Information System (INIS)

    Yoshitomi, H.; Kowatari, M.

    2016-01-01

    Both a cylinder and a slab phantom have been recommended to be used as calibration phantoms for eye lens dosimetry in the International Atomic Energy Agency TECDOC. This study describes investigations on the influence of the type of phantom on the calibration of dosemeters. In order to fulfil the purpose, backscatter radiation from practically used water-filled phantoms was evaluated by calculations and experiments. For photons, the calculations showed that the cylinder phantom had 10 % lower backscattered effect at maximum than a slab phantom, and simulated well the backscattered effect of the human head or neck to within ±10 %. The irradiation results of non-filtered optically stimulated luminescence and radio-photoluminescence glass dosemeters indicated that the differences of the calibration factors between the two types of phantoms were up to 20 and 10 %, respectively, reflecting the response to backscattered photons. For electrons, no difference was found between the two types of phantoms. (authors)

  13. Dosimetric characteristics of a TLD dosemeter with extremities

    International Nuclear Information System (INIS)

    Molina P, D.; Diaz B, E.; Lien V, R.

    1999-01-01

    It was designed a TLD dosemeter for the monitoring of the extremities. This one consists in a metallic ring with a circular orifice where is arranged a T L detector of LiF: Mg,Ti (Model JR1152C) 5 x 5 x 0.8 mm 3 covered by a polyethylene fine layer. In this work were studied the dosimetric properties of the dosemeter for its application in the dosimetry of extremities for photonic radiation. the results obtained allow conclude that the designed dosemeter can be used for the extremities monitoring. (Author)

  14. A personal radon dosemeter

    International Nuclear Information System (INIS)

    Oberstedt, S.; Vanmarcke, H.

    1994-03-01

    In the last decade the radon issue has become one of the major problems of radiation protection. Animal studies as well as epidemiological studies showed an increased lung cancer risk. A new personal radon-dosemeter on the basis of a CR-39 (poly-allyl diglycol carbonate) track-etch detector has been developed. The read-out of the detectors is based on the image- processing technique. The actual efficiency of the new dosemeter, obtained with a semi-automatic personal-computer based image-analysis system, is 1.43 +/- 0.15 tracks/cm 2 /(kBq/m 3 h), which is about three times that of the widely used Karlsruhe-type detector based on polycarbonate detectors

  15. Using operational equipment to read accident dosemeters.

    Science.gov (United States)

    Devine, R T; Vigil, M M; Martinez, W A

    2004-01-01

    Analysis of accident dosemeters usually involves the use of laboratory-based counting equipment. Gamma spectrometers are used for indium, copper and gold, and alpha-beta detectors for sulphur. This equipment is usually not easily transported due to the shielding required and the weight and delicacy of the counters. For intercomparison studies that require reading the dosemeters on site, a transportable system is required unless the site operating the study can count samples for all the participants. In the case of an actual accident these systems would have a difficulty in counting a large number of accident dosemeters. In an accident, personnel are usually subdivided according to their level of exposure. Those exposed to higher doses are treated immediately. An alternate system should be made available to handle the dosemeters worn by those personnel are likely to receive lower doses. Improvements in portable operational equipment for gamma and beta monitoring allow their use as spectrometers. Such a system was used for the SILENE intercomparison conducted at IRSN Valduc on 12 June and 19, 2002, and the preliminary results compared well with the other participants.

  16. Calibration of clinical dosemeters in the IAEA water phantom

    International Nuclear Information System (INIS)

    Caldas, L.V.E.; Albuquerque, M.P.P.

    1994-01-01

    The procedures recommended by the IAEA Code of Practice were applied at the Calibration Laboratory of Sao Paulo in order to provide in the future the clinical dosemeters users with absorbed dose to water calibration factors for Cobalt 60 radiation beams. In this work the clinical dosemeters were calibrated free in air and in water, and the results were compared, using conversion factors. The several tested clinical dosemeters of different manufacturers and models belong to the laboratory and to hospitals. For the measurements in water the IAEA cubic water phantom was used. The dosemeters were all calibrated free in air in terms of air kerma, and the calibration factors in terms of absorbed dose to water were obtained through conversion factors. the same dosemeters were also calibrated into the water phantom. Good agreement was found between the two methods, the differences were always less than 0.5%. The data obtained during this work show that when the dosemeters are used only in Cobalt 60 radiation and the users apply in the hospital routine work the IAEA Code of Practice, the calibration can be performed directly in the water phantom. This procedure provides the useful calibration factors in terms of absorbed dose to water

  17. Photon beam audits for radiation therapy clinics: A pilot mailed dosemeter study in Turkey

    International Nuclear Information System (INIS)

    Yegingil, Z.; DeWerd, L. A.; Davis, S. D.; Hammer, C.; Kunugi, K.

    2012-01-01

    A thermoluminescent dosemeter (TLD) mailed dose audit programme was performed at five radiotherapy clinics in Turkey. The intercomparison was organised by the Univ. of Wisconsin Radiation Calibration Laboratory (UWRCL), which was responsible for the technical aspects of the study including reference irradiations, distribution, collection and evaluation. The purpose of these audits was to perform an independent dosimetry check of the radiation beams using TLDs sent by mail. Acrylic holders, each with five TLD chips inside and instructions for their irradiation to specified absorbed dose to water of 2 Gy, were mailed to all participating clinics. TLD irradiations were performed with a 6 MV linear accelerator and 60 Co photon beams. The deviations from the TL readings of UWRCL were calculated. Discrepancies inside the limits of ±5 % between the participant-stated dose, and the TLD-measured dose were considered acceptable. One out of 10 beams checked was outside this limit, with a difference of 5.8 %. (authors)

  18. Evaluation of the thermoluminescent detector answers of CaSO{sub 4}:Dy, LiF:Mg,Ti and micro LiF:Mg,Ti in photon clinical beams dosimetry using water simulator; Avaliacao da resposta de detectores termoluminescentes de CaSO4:Dy, LiF:Mg,Ti e microLiF:Mg,Ti na dosimetria de feixes clinicos de fotons utilizando simulador de agua

    Energy Technology Data Exchange (ETDEWEB)

    Matsushima, Luciana C.; Veneziani, Glauco R.; Campos, Leticia L., E-mail: lmatsushima@usp.b, E-mail: veneziani@ipen.b, E-mail: lcrodri@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (GMR/IPEN/CNEN-SP), Sao Paulo, SP (Brazil). Gerencia de Metrologia das Radiacoes; Sakuraba, Roberto K.; Cruz, Jose C. da, E-mail: rsakuraba@einstein.b, E-mail: jccruz@einstein.b [Sociedade Beneficente Israelita Brasileira, Sao Paulo, SP (Brazil). Hospital Albert Einstein (HAE)

    2011-10-26

    This paper perform the comparative study of thermoluminescent answer of calcium sulfate dosemeter doped with dysprosium (DaSO{sub 4}:Dy) produced by the IPEN, Sao Paulo, with answer of lithium fluoride dosemeters doped with magnesium and titanium (LiF:Mg, Ti) in the dosimetry of clinical beams of photons (6 and 15 MV) by using water simulator object. Dose-answer curves were obtained for gamma radiation of cobalt-60 in the air and in conditions of electronic equilibrium (plate of PMMA), and clinical photons of CLINAC model 2100C accelerators of the two evaluated hospitals: Hospital das Clinicas of the Faculty of Medicine of Sao Paulo university and Hospital Albert Einstein. It was also evaluated the sensitivity and reproduction of the three dosemeters

  19. Performance of CVD diamond as an optically and thermally stimulated luminescence dosemeter

    International Nuclear Information System (INIS)

    Preciado-Flores, S.; Schreck, M.; Melendrez, R.; Chernov, V.; Bernal, R.; Cruz-Vazquez, C.; Cruz-Zaragoza, E.; Barboza-Flores, M.

    2006-01-01

    Diamond is a material with extreme physical properties. Its radiation hardness, chemical inertness and tissue equivalence qualify it as an ideal material for radiation dosimetry. In the present work, the optically stimulated luminescence (OSL) and thermoluminescence (TL) characteristics of a 10 μm thick CVD diamond (polycrystalline diamond films prepared by chemical vapor deposition) film were studied in order to test its performance as a beta radiation dosemeter. The TL response is composed of four main TL glow peaks; two of these are in the range of 150-200 deg. C and two additional peaks in the 250-400 deg. C temperature range. The integrated TL as a function of radiation dose is linear up to 100 Gy and increases with increasing dose exposure. The dose dependence of the integrated OSL exhibits a similar behavior. The observed OSL/TL behavior for the CVD diamond film clearly demonstrate its capability for applications in radiation dosimetry with special relevance in medical dosimetry owing to the diamond's intrinsic material properties. (authors)

  20. Elaboration And Study Of Transfer Alanine/ESR Dosemeters

    International Nuclear Information System (INIS)

    Torres G, Luz A.

    1996-01-01

    The dosimetry is the dose measure imparted by the energy from the ionizing radiation to the matter. The dosemeters is the means used for the determination of such a dose. Diverse dosimetric classes exist, this classification depends in essence of the energy involved in the irradiation process and of its application necessity. It is as well as in radiological protection movie dosemeter is used, and TLD, in the calibration of units, as those of cobalt, the ionization cameras are used, in the detection of superficial contamination the accountants Geiger Muller and proportional etc. The transfer dosimeter Alanine/ESR is used, object of the present work, is characterized because after reading the registered dose, the information is conserved and it can be transferred and read in any team of ESR. Likewise the following dose that is imparted will be added to the previous one registered, this indicates that the dosemeters is of integral character. In the spectra taken ESR, it is determined that the integral double or area under the curve of the spectrum ESR is proportional to the concentration of free radicals generated by the radiation ionization and in turn this concentration is proportional to the dose received by the pills; in last, these proportionality relationships take to that the area under the curve of the spectrum main ESR is proportional to the radiation dose received by the alanine dosemeters. This dosemeter seeks to be an economic alternative that it will lend the calibration service that today offers the ionization Cameras or the acrylic red industrial dosemeters

  1. Australian participation in international intercomparisons of environmental dosemeters

    International Nuclear Information System (INIS)

    Boas, J.F.; David, R.C.; Jones, J.

    2002-01-01

    The Environmental Measurements Laboratory of the U.S. Department of Energy has conducted international intercomparisons of environmental dosemeters at regular intervals since 1974. This paper reports the results obtained in the 10th and 11th intercomparisons by JFB and JGY at the Australian Radiation Laboratory (in 1992 and 1996 respectively) and in the 12th intercomparison by JFB and DRCT at the Australian Radiation Protection and Nuclear Safety Agency and by JGY and JJ at Australian Radiation Serviceces during 2000 and 2001. The ARL/ARPANSA measurements used a simple dosemeter holder containing two CaSO 4 :Dy impregnated Teflon discs, one unshielded and the other shielded by approximately 2.4 mm of Cu. The ARS measurements used a commercially available TLD badge case containing a CaSO 4 :Dy impregnated Teflon card. The commercial badge case was not symmetrical in design. The filtration in the front and rear sections of the commercial badge case varied from an open area to an area shielded by copper, tin and cadmium. The results obtained were generally within ±15% of the reference values measured by EML using high-pressure ionisation chambers for dosemeters exposed hi the field to environmental radiation and for dosemeters given a single exposure to 137 Cs or 241 Am at EML or an associated laboratory. Similar agreement was obtained for dosemeters exposed in the field and also given a single exposure to 137 Cs radiation. In order to obtain this measure of agreement, the dosemeters were individually calibrated and the readouts corrected for fading. Copyright (2002) Australasian Radiation Protection Society Inc

  2. Thermoluminescent characterization of thin films of aluminium oxide submitted to beta and gamma radiation; Caracterizacion termoluminiscente de peliculas delgadas de oxido de aluminio sometidas a radiacion beta y gamma

    Energy Technology Data Exchange (ETDEWEB)

    Villagran, E.; Escobar A, L.; Camps, E.; Gonzalez, P.R.; Martinez A, L. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2002-07-01

    By mean of the laser ablation technique, thin films of aluminium oxide have been deposited on kapton substrates. These films present thermoluminescent response (Tl) when they are exposed to beta and gamma radiation. The brilliance curves show two peaks between 112 C and 180 C. A dose-response relationship study was realized and the Tl kinetic parameters were determined using the computerized deconvolution of the brilliance curve (CGCD). The thin films of aluminium oxide have potential applications as ultra.thin radiation dosemeters. (Author)

  3. The thermoluminescence as tool in the reconstruction of volcanic events

    International Nuclear Information System (INIS)

    Ramirez L, A.; Schaaf, P.; Martin del Pozzo, A.L.; Gonzalez M, P.

    2000-01-01

    Within the Mexican land a great number of volcanoes are situated which a considerable part of them are still active. The relevance of dating pomex deposits, ash or lava of these poly genetic volcanoes is to determine the periodicity and magnitude of the volcanic events happened. In this work is presented the preliminary result of the dating by thermoluminescence in a pomex of a pyroclastic flux coming from a volcano in the state of Puebla with the purpose of providing elements to the knowledge which describe the eruptive history of the explosive volcanism at center of Mexico. For the sample dating the volcanic glasses of pomex were separated and it was applied the fine grain technique with a grain size between 4-11 μ m. In order to calculate the rate of annual dose it was carried out the following: in the determination of 238 U and 232 Th radioisotope concentration was used the neutron activation technique in a nuclear reactor, in the determination of the K 40 radioisotope was used a scanning electron microscope, the rate of environmental and cosmic dose was measured arranging Tl dosemeters of CaSO 4 : Dy in the sampling place. In order to calculate the paleodoses it was carried out the following: the equivalent dose (Q) was determined starting form the additive method and the supra linearity factor (I) starting from regenerative method and in both methods the irradiated process was realized with a 90 Sr beta source. With the above determinations it was calculated a paleodoses of 231 Gy and a rate of annual dose of 6.074 x 10 -3 Gy/year, estimating an age of: Age pomez = 231 Gy / 6.074 Gy x 10 -3 Gy /year = 38030 ± 4000 years. (Author)

  4. Improvement of the calibration technique of clinical dosemeters

    International Nuclear Information System (INIS)

    Ehlin Caldas, L.V.

    1988-08-01

    Clinical dosemeters constituted of ionization chambers connected to electrometers are usually calibrated as whole systems in appropriate radiation fields against secondary standard dosemeters in calibration laboratories. This work reports on a technique of component calibration procedures separately for chambers and electrometers applied in the calibration laboratory of IPEN-CNEN, Brazil. For electrometer calibration, redundancy was established by using a standard capacitor of 1000pF (General Radio, USA) and a standard current source based on air ionization with Sr 90 (PTW, Germany). The results from both methods applied to several electrometers of clinical dosemeters agreed within 0.4%. The calibration factors for the respective chambers were determined by intercomparing their response to the response of a certified calibrated chamber in a Co 60 calibration beam using a Keithley electrometer type 617. Overall calibration factors compared with the product of the respective component calibration factors for the tested dosemeters showed an agreement better than 0.7%. This deviation has to be considered with regard to an uncertainty of 2.5% in routine calibration of clinical dosemeters. Calibration by components permits to calibrate ionization chambers one at a time for those hospitals who have several ionization chambers but only one electrometer (small hospitals, hospitals in developing countries). 6 refs, 2 figs, 2 tabs

  5. Specification for personal photographic dosemeters

    International Nuclear Information System (INIS)

    1981-01-01

    The 1981 British/International Standard Specification, prepared under the direction of the Nuclear Engineering Standards Committee and TC85 of the International Organization for Standardization, is described for personal photographic dosemeters. The Standard specifies classification, characteristics and test procedures to determine absorbed doses due to X or gamma radiations (energy less than 3 MeV) and absorbed doses due to beta radiation (max. energy 0.6 to 3 MeV), whether or not accompanied by X, gamma or bremsstrahlung photon radiation. The Standard is particularly applicable to dosemeters intended to be carried on the chest or wrist. (U.K.)

  6. Handbook of thermoluminescence

    CERN Document Server

    Furetta, Claudio

    2010-01-01

    This second edition of the Handbook of Thermoluminescence enlarges on all the subjects which were treated in the first edition and adds further arguments, including the theory of thermoluminescent dose measurement, several examples concerning the kinetics parameters determination using various methods such as peak shape, isothermal decay, and so on. A special section is devoted to food irradiation, an important subject at the present time, and to the thermoluminescent characterization of the minerals extracted from the irradiated food. Another new section is devoted to the thermoluminescent ph

  7. Calibration of Far West Technology (FWT-60) radiachromic dye dosemeters

    International Nuclear Information System (INIS)

    Mincher, B.J.; Zaidi, M.K.

    1992-01-01

    Radiachromic dye dosimetry was used to measure kilogray doses absorbed by various liquid samples during gamma-ray exposure in a spent nuclear fuel pool. The source of nuclear fuel was the Advanced Test Reactor (ATR) at the Idaho National Engineering Laboratory (INEL). Calibrations were performed using a 60 Co source and were run on bare dosemeters, as well as on dosemeters which mocked to simulate the samples. These dosemeters were prepared as a dye-impregnated nylon film and are commercially available. When exposed to gamma-ray doses of 0.5 to 200 kGy, a color change occurs which has an optical density proportional to absorbed dose. The difference in the calibration curves demonstrated the importance of irradiation of dosemeters under conditions as close to the actual samples as possible. Since these dosemeters could not be immersed directly in the organic solutions of interest, they were sandwiched between layers of lucite and stainless steel. This simulated the conditions inside an irradiated sample and provides a practical method of measuring absorbed doses. The reproducibility of measurements using the radiachromic dye dosemeters is also shown. 10 refs

  8. Absorbed dose measurements in mammography using Monte Carlo method and ZrO2+PTFE dosemeters

    International Nuclear Information System (INIS)

    Duran M, H. A.; Hernandez O, M.; Salas L, M. A.; Hernandez D, V. M.; Vega C, H. R.; Pinedo S, A.; Ventura M, J.; Chacon, F.; Rivera M, T.

    2009-10-01

    Mammography test is a central tool for breast cancer diagnostic. In addition, programs are conducted periodically to detect the asymptomatic women in certain age groups; these programs have shown a reduction on breast cancer mortality. Early detection of breast cancer is achieved through a mammography, which contrasts the glandular and adipose tissue with a probable calcification. The parameters used for mammography are based on the thickness and density of the breast, their values depend on the voltage, current, focal spot and anode-filter combination. To achieve an image clear and a minimum dose must be chosen appropriate irradiation conditions. Risk associated with mammography should not be ignored. This study was performed in the General Hospital No. 1 IMSS in Zacatecas. Was used a glucose phantom and measured air Kerma at the entrance of the breast that was calculated using Monte Carlo methods and ZrO 2 +PTFE thermoluminescent dosemeters, this calculation was completed with calculating the absorbed dose. (author)

  9. Summary of personal neutron dosemeter results obtained within the EVIDOS project

    International Nuclear Information System (INIS)

    Luszik-Bhadra, M.; Reginatto, M.; Schuhmacher, H.; Bolognese-Milsztajn, T.; Lacoste, V.; Boschung, M.; Fiechtner, A.; Coeck, M.; Vanhavere, F.; Curzio, G.; Errico d', F.; Kylloenen, J.-E.; Lindborg, L.; Bartlett, D.; Tanner, R.

    2005-01-01

    Full text: Within the EC project EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields'), different types of neutron personal dosemeters were irradiated in workplace fields in nuclear installations in Europe. Mostly electronic personal neutron dosemeters were tested, among them commercial devices (Thermo Electron EPD-N, Aloka PDM-313), devices from first industrial prototype series (Thermo Electron EPD-N2, Saphydose-n) and laboratory prototypes which were already in the stage of lightweight battery-operated instruments (PTB DOS-2002). In addition, dosemeters with (almost) immediate readout (BTI bubble detectors, Rados DIS-N) and passive dosemeters which needed no fielddependent calibration factors (PADC track detectors from PSI and NRPB) were used, as well as those TLD albedo dosemeters which are routinely used in the facilities visited. The results of measurements obtained within the EVIDOS project in workplace fields in nuclear installations in Europe, i.e. at Kruemmel (boiling water reactor, transport cask), at Mol (Venus research reactor, fuel facility Belgonucleaire) and at Ringhals (pressurized water reactor, transport cask) are presented and compared to reference values of personal dose equivalent H p (10) determined by means of Bonner spheres and novel directional spectrometers. In fields with strong back-scatter and moderation of neutrons, several dosemeters showed overreadings by more than a factor of two. In strongly directed fields with neutron dose equivalent peaked at about 1 MeV (MOX fuel), the same dosemeters showed under-readings of about a factor of two. These under- and over-readings in the workplace fields can be explained in terms of the dosemeter response functions for mono-energetic neutrons, which show over-responses in the thermal and intermediate neutron energy region and under-responses at about 1 MeV neutron energy. The dosemeter readings obtained in the workplace fields were checked for consistency by

  10. The evaluation of multi-element personal dosemeters using the linear programming method

    International Nuclear Information System (INIS)

    Kragh, P.; Ambrosi, P.; Boehm, J.; Hilgers, G.

    1996-01-01

    Multi-element dosemeters are frequently used in individual monitoring. Each element can be regarded as an individual dosemeter with its own individual dose measurement value. In general, the individual dose values of one dosemeter vary according to the exposure conditions, i. e. the energy and angle of incidence of the radiation. The (final) dose measurement value of the personal dosemeter is calculated from the individual dose values by means of an evaluation algorithm. The best possible dose value, i.e. that of the smallest systematic (type B) uncertainty if the exposure conditions are changed in the dosemeter's rated range of use, is obtained by the method of linear programming. (author)

  11. Thermoluminescence as a probe in bioactivity studies; the case of 58S sol-gel bioactive glass

    International Nuclear Information System (INIS)

    Polymeris, George S; Tsirliganis, Nestor C; Goudouri, Ourania Menti; Paraskevopoulos, Konstantinos M; Kontonasaki, Eleana; Kitis, George

    2011-01-01

    The formation of a carbonated hydroxyapatite (HCAp) layer on the surface of bioactive materials is the main reaction that takes place upon their immersion in physiological fluids. To date, all techniques used for the identification of this HCAp formation are rather time consuming and not well suited to detailed and rapid monitoring of changes in the bioactivity response of the material. The aim of this work is to explore the possibility of using thermoluminescence (TL) for the discrimination between different bioactive responses in the case of the 58S bioactive glass. Results provided strong indications that the 110 deg. C TL peak of quartz can be used effectively in the study of the bioactive behaviour of 58S bioactive glass, since it is unambiguously present in all samples and does not require deconvolution analysis. Furthermore, the intensity of the 110 deg. C TL peak is proven to be very sensitive to the different bioactive responses, identifying the loss of silica which takes place at the first stages of the sequence. The discontinuities of the 110 deg. C TL peak intensity plot versus immersion time at 8 and 1440 min provide experimental indications regarding the timescale for both the beginning of amorphous CaP formation as well as the end of crystalline hydroxyl-apatite formation respectively, while the spike in the sensitization of the 110 deg. C TL peak, which was observed for immersion times ranging between 20 and 40 min, could be an experimental feature indicating the beginning of the crystalline HCAp formation.

  12. Environmental radiation measurements using TLD in and around AERE, Dhaka

    International Nuclear Information System (INIS)

    Mollah, A.S.; Husain, S.R.; Rahman, M.M.

    1986-01-01

    The external background radiation level in and around the Atomic Energy Research Establishment (AERE) in Dhaka has been measured. The measurements were performed using lithium fluoride (LiF) thermoluminescence dosemeters (TLD-100) at 32 locations, all one metre above the ground. The annual average dose rate measured in the AERE environs was 1.74+-0.23 mGy.y -1 in air, based on analysis of thermoluminescence dosemeter data collected from 1982 to 1984. (author)

  13. Determination of isodose curves in Radiotherapy using an Alanine/ESR dosemeter

    International Nuclear Information System (INIS)

    Chen, F.; Baffa, O.; Graeff, C.F.O.

    1998-01-01

    It was studied the possible use of an Alanine/ESR dosemeter in the isodose curves mapping in normal treatments of Radiotherapy. It was manufactured a lot of 150 dosemeters with base in a mixture of D-L Alanine dust (80 %) and paraffin (20 %). Each dosemeter has 4.7 mm diameter and 12 mm length. A group of 100 dosemeters of the lot were arranged inside 50 holes of the slice 25 of the phantom Rando Man. The phantom irradiation was realized in two opposed projections (AP and PA) in Co-60 equipment. A group of 15 dosemeters was take of the same lot for obtaining the calibration curve in a 1-20 Gy range. After irradiation the signal of each dosemeter was measured in an ESR spectrometer operating in the X-band (∼ 9.5 GHz) and the wideness of Alanine ESR spectra central line was correlated with the radiation dose. The wideness dose calibration curve resulted linear with a correlation coefficient 0.9996. The isodose curves obtained show a profile enough similar at comparing with the theoretical curves. (Author)

  14. Individual monitoring with official electronic dosemeters in Germany - concepts and reality

    International Nuclear Information System (INIS)

    Martini, E.; Wahl, W.; Huebner, S.; Freynhagen, D.; Staemmler, M.

    2005-01-01

    Full text: Active (electronic) personal dosemeters (APD) are radiometers, which are equipped as active or passive dosimeters. They can directly indicate a dose rate and/or the dose integrated during a certain period (direct-indicating, active electronic personal dosemeters mentioned) or only over a separate reader (direct-selectable, passive electronic personal dosemeters mentioned). A concept is presented for the use of passive or active electronic personal dosemeters in Germany as official dosemeters, e.g. in hospitals. The concept features a net-based approach for secure data communication between readers for passive and active electronic personal dosemeters und the official personal dosimetry monitoring service(s), taking the German policies (StrISchV und RoeV in hospitals) into account. The net-based solution includes a processor controlled interface via TCP/IP connected to the dosimetry reader(s), und reader interfaces, an official interim data bank and all the necessary equipment such as PC, Raid-configuration, USV support, network connection and further details. Investigations have been started to achieve highest-level data manipulation security, data completeness und data correctness. (author)

  15. Understanding differences in dose-equivalents reported by passive and electronic personal dosemeters

    International Nuclear Information System (INIS)

    Perks, Christopher A.; Burgess, Peter; Smith, David; Salasky, Mark; Yahnke, Cliff

    2008-01-01

    Full text: In a number of challenging environments, clients occasionally double badge with electronic personal dosemeters (EPDs) to ensure day-to-day management of their employees personal dose-equivalent while using passive (in our case Luxel or InLight) dosemeters for monthly monitoring for approved results for dose record keeping. In some cases there have been significant differences in the cumulative doses recorded by the EPDs and the passive dosemeters. In these circumstances the passive dosemeters usually report a higher dose than the EPD by up to a factor of two, though more commonly 1.3. In this paper we describe the differences seen between EPD and passive dosemeters (in a number of countries). We then examine the possible causes for such discrepancies by comparison with published response function data available for the EPDs and Landauer dosemeters. We have undertaken a number of experiments comparing directly a number of EPDs and passive dosemeter response to a variety of energy and complex angular geometries where the two types of dosimeter have been exposed at the same time. Recommendations will be made on the appropriate use of double badging in difficult environments and interpretation of the results. (author)

  16. Microcomputer-controlled thermoluminescent analyser IJS MR-200; Mikroracunalniski termoluminescentni analizator IJS MR-200

    Energy Technology Data Exchange (ETDEWEB)

    Mihelic, M; Miklavzic, U; Rupnik, Z; Satalic, P; Spreizer, F; Zerovnik, I [Institut Jozef Stefan, Ljubljana (Yugoslavia)

    1985-07-01

    Performances and concept of the multipurpose, microcomputer-controlled thermoluminescent analyser, designed for use in laboratory work TL dosemeters as well as for routine dose readings in the range from ecological to accident doses is described. The main features of the analyser are: time-linear sampling, digitalisation, storing, and subsequent displaying on the monitor time scale of the glow and and temperature curve of the TL material; digital stabilization, control and diagnostic of the analog unit; ability of storing 7 different 8-parametric heating programs; ability of storing 15 evaluation programs defined by 2 or 4 parameters and 3 different algorithms (altogether 5 types of evaluations). Analyser has several features intended for routine work: 9 function keys and possibilities of file forming on cassette or display disc, of dose calculation and averaging, of printing reports with names, and possibility of additional programming in Basic. (author)

  17. Using signal ''KVANT-1'' direct-reading dosemeter for the purposes of personnel monitoring

    International Nuclear Information System (INIS)

    Glinskij, G.A.; Karasev, V.S.; Mukhin, I.E.; Chumak, V.K.

    1977-01-01

    Presented is the description of ''KVANT-1'' dosemeter for monitoring personnel doses of gamma and X radiation. The dosemeter permits to judge on the radiation intensity, to control directly the dose being accumulated, to store the reading of the dose accumulated for a necessary period of time, to obtain sound signal in case of reaching the limit of a pre-set dose. Presented are a general view, block diagram, and the discription of the dosemeter desing and operation. Advantages of the ''KVANT-1'' dosemeter are shown as compared with the conventional personnel monitoring IFK-2,3 and KID-2 dosemeters [ru

  18. CVD diamonds as thermoluminescent detectors for medical applications

    International Nuclear Information System (INIS)

    Marczewska, B.; Olko, P.; Nesladek, M.; Waligorski, M.P.R.; Kerremans, Y.

    2002-01-01

    Diamond is believed to be a promising material for medical dosimetry due to its tissue equivalence, mechanical and radiation hardness, and lack of solubility in water or in disinfecting agents. A number of diamond samples, obtained under different growth conditions at Limburg University, using the chemical vapour deposition (CVD) technique, was tested as thermoluminescence dosemeters. Their TL glow curve, TL response after doses of gamma rays, fading, and so on were studied at dose levels and for radiation modalities typical for radiotherapy. The investigated CVD diamonds displayed sensitivity comparable with that of MTS-N (Li:Mg,Ti) detectors, signal stability (reproducibility after several readouts) below 10% (1 SD) and no fading was found four days after irradiation. A dedicated CVD diamond plate was grown, cut into 20 detector chips (3x3x0.5 mm) and used for measuring the dose-depth distribution at different depths in a water phantom, for 60 Co and six MV X ray radiotherapy beams. Due to the sensitivity of diamond to ambient light, it was difficult to achieve reproducibility comparable with that of standard LiF detectors. (author)

  19. A CR-39 track dosemeter for routine individual neutron monitoring

    International Nuclear Information System (INIS)

    Luszik-Bhadra, M.; Alberts, W.G.; Dietz, E.; Guldbakke, S.; Matzke, M.; d'Errico, F.

    1994-01-01

    A personal neutron dosemeter for routine individual monitoring is proposed. It is based on a CR-39 track detector covered on three separate areas by converters with different boron contents and inserted into a commercial TLD albedo dosemeter capsule. The dose readings from three areas of the electrochemically etched CR-39 detector are combined to yield a dose equivalent response which is almost independent of the incident neutron energy in the range from thermal neutrons up to 20 MeV. In addition, the dose contributions of thermal, intermediate and fast neutrons can be determined separately. Unlike the TLD albedo dosemeter, which in general requires prior in-field calibration and whose use is then restricted to that field, this dosemeter can be used in neutron fields without any knowledge of the spectral distribution with the same calibration factor. The angular dependence of the dosemeter's response has been measured and compared with that of the directional dose equivalent H'(10). The lower limit of detection is 0.15 mSv. It is possible to obtain an independent, second dose reading from the same Cr-39 detector for neutron energies above 100 keV. The dosemeter has also been successfully tested for use in accident dosimetry applying chemical etching and an optical density reading of the CR-39 detector. (author)

  20. Calibration of farmer dosemeters

    International Nuclear Information System (INIS)

    Ahmad, S.S.; Anwar, K.; Arshed, W.; Mubarak, M.A.; Orfi, S.D.

    1984-08-01

    The Farmer Dosemeters of Atomic Energy Medical Centre (AEMC) Jamshoro were calibrated in the Secondary Standard Dosimetry Laboratory (SSDL) at PINSTECH, using the NPL Secondary Standard Therapy level X-ray exposure meter. The results are presented in this report. (authors)

  1. Passive dosemeter

    International Nuclear Information System (INIS)

    Hassib, G.; Piesch, E.

    1981-01-01

    The dosemeter will be used for the detection of radon and thoron as well as their daughter products. It consists of two chambers with nuclear trace or alpha detectors. Both chambers are connected with one another at their front sides with a filter arranged in between. One of the chambers has got lateral openings, it is used as a diffusion chamber. In the other chamber exclusively radon and thoron are detected. (DG) [de

  2. Measurement of TLD Albedo response on various calibration phantoms

    International Nuclear Information System (INIS)

    Momose, T.; Tsujimura, N.; Shinohara, K.; Ishiguro, H.; Nakamura, T.

    1996-01-01

    The International Commission on Radiation Units and Measurements (ICRU) has recommended that individual dosemeter should be calibrated on a suitable phantom and has pointed out that the calibration factor of a neutron dosemeter is strongly influenced by the the exact size and shape of the body and the phantom to which the dosemeter is attached. As the principle of an albedo type thermoluminescent personal dosemeter (albedo TLD) is essentially based on a detection of scattered and moderated neutron from a human body, the sensitivity of albedo TLD is strongly influenced by the incident neutron energy and the calibration phantom. (1) Therefore for albedo type thermoluminescent personal dosemeter (albedo TLD), the information of neutron albedo response on the calibration phantom is important for appropriate dose estimation. In order to investigate the effect of phantom type on the reading of the albedo TLD, measurement of the TLD energy response and angular response on some typical calibration phantoms was performed using dynamitron accelerator and 252 Cf neutron source. (author)

  3. Thermoluminescent dosimetry in pediatric patients subjected to TAC cranium multi cutting studies

    International Nuclear Information System (INIS)

    Tabares V, M.J.; Azorin N, J.; Azorin V, J.C.

    2007-01-01

    Full text: In this work the doses to critical organs in pediatric patient subjected to studies of multi cutting skull tomography were determined. The study included patient smaller than 16. The used tomograph was a Siemens Somaton 16 plus multi cutting. The dosimetry one carries out with thermoluminescent dosemeters of LiF:Mg,Cu,P +Ptfe of national production. This dosemeters type was used due to it characteristics, such as low fading, equivalence with the tissue, easy handling and reading and independence of it answer with the radiation energy. The value of the absorbed doses in organs in the simple study results half of the obtained value in the contrasted study that which indicates that the CTDI vol is proportional to the dose to organ. In the case of the skull three-dimensional reconstruction tomography the CTDI vol was practically the same one that for simple tomography was practically; however, the doses to organs differed significantly. When comparing the TDI vol value of the three-dimensional reconstruction tomography with that of the contrasted tomography the obtained reason it was 1:2; however, the doses to organ didn't show the same relationship and the absorbed dose in the case of the thyroid gland was even bigger in the study of three-dimensional reconstruction. The value of the CTDI vol is a good indicator of the exposure factors selected for the tomographic exploration and it is useful for the determination of the effective dose, but it doesn't indicate the absorbed dose to organ. (Author)

  4. Chemical dosimetry system for criticality accidents.

    Science.gov (United States)

    Miljanić, Saveta; Ilijas, Boris

    2004-01-01

    Ruder Bosković Institute (RBI) criticality dosimetry system consists of a chemical dosimetry system for measuring the total (neutron + gamma) dose, and a thermoluminescent (TL) dosimetry system for a separate determination of the gamma ray component. The use of the chemical dosemeter solution chlorobenzene-ethanol-trimethylpentane (CET) is based on the radiolytic formation of hydrochloric acid, which protonates a pH indicator, thymolsulphonphthalein. The high molar absorptivity of its red form at 552 nm is responsible for a high sensitivity of the system: doses in the range 0.2-15 Gy can be measured. The dosemeter has been designed as a glass ampoule filled with the CET solution and inserted into a pen-shaped plastic holder. For dose determinations, a newly constructed optoelectronic reader has been used. The RBI team took part in the International Intercomparison of Criticality Accident Dosimetry Systems at the SILENE Reactor, Valduc, June 2002, with the CET dosimetry system. For gamma ray dose determination TLD-700 TL detectors were used. The results obtained with CET dosemeter show very good agreement with the reference values.

  5. Dosimetric properties of dysprosium doped lithium borate glass irradiated by 6 MV photons

    Science.gov (United States)

    Ab Rasid, A.; Wagiran, H.; Hashim, S.; Ibrahim, Z.; Ali, H.

    2015-07-01

    Undoped and dysprosium doped lithium borate glass system with empirical formula (70-x) B2O3-30 Li2O-(x) Dy2O3 (x=0.1, 0.3, 0.5, 0.7, 1.0 mol%) were prepared using the melt-quenching technique. The dosimetric measurements were performed by irradiating the samples to 6 MV photon beam using linear accelerator (LINAC) over a dose range of 0.5-5.0 Gy. The glass series of dysprosium doped lithium borate glass produced the best thermoluminescence (TL) glow curve with the highest intensity peak from sample with 1.0 mol% Dy2O3 concentration. Minimum detectable dose was detected at 2.24 mGy, good linearity of regression coefficient, high reproducibility and high sensitivity compared to the undoped glass are from 1.0 mol% dysprosium doped lithium borate glass. The results indicated that the series of dysprosium doped lithium glasses have a great potential to be considered as a thermoluminescence dosimetry (TLD).

  6. Evaluation of film dosemeters by linear programming

    International Nuclear Information System (INIS)

    Kragh, P.; Nitschke, J.

    1992-01-01

    An evaluation method for multi-component dosemeters is described which uses linear programming in order to decrease the dependence on energy and direction. The results of this method are more accurate than those obtained with the evaluation methods so far applied in film dosimetry. In addition, systematic errors can be given when evaluating individual measurements. Combined linear programming, as a special case of the presented method, is described taking a film dosemeter of particular type as an example. (orig.) [de

  7. Dosemeter calibration in X-ray and in cobalt-60

    International Nuclear Information System (INIS)

    Silva, T.A. da

    1988-01-01

    Some tests about quality security for clinical dosemeter calibration in secondary standard dosimetry laboratory are described. The tests in gama calibration system, in X-ray calibration, in secondary standard dosimeter, in the dosemeter that will be calibrated, during the calibration and after the calibration are shown. (C.G.C.) [pt

  8. Extremity dosimetry in nuclear medicine services using thermoluminescent detectors

    International Nuclear Information System (INIS)

    Mebhah, D.; Djeffal, S.; Badreddine, A.; Medjahed, M.

    1993-01-01

    The Radiation Protection and Safety Centre in Algiers provides two types of dosemeters, one for monitoring doses to the whole body and skin and the other one for monitoring doses to the extremities of the body. In nuclear medicine services and radiopharmaceutical laboratories, hands and arms are often closer to a given radiation source than the main part of the body and therefore receive greater doses. In this context, extremity doses have been measured by a ring dosemeter and by a fingertip ultra-thin dosemeter. The ring dosemeter consists of a metallic ring with a circular indentation to hold a LiF chip which is covered with a 10 mg.cm -2 shrinkable black polyamide layer. The ultra-thin dosemeter contains a 5 mg.cm -2 LiF element for measuring doses at a depth of 7 mg.cm -2 . These extremity dosemeters have been characterised before their use in the field. They have also been tested using radioisotopes of various energies. The doses received by the monitored workers were correlated with the amount of the handled activity. The doses obtained using the fingertip and the ring dosemeters are presented and discussed from a radiological point of view. (author)

  9. Monte Carlo verification of control-rod worth for the Savannah River K reactor

    International Nuclear Information System (INIS)

    Mosteller, R.D.

    1992-01-01

    The Savannah River K Reactor is a heavy-water reactor that relies on control-rod movement to control its reactivity and power distribution during normal operations. It is necessary, therefore, to have an accurate estimate of the reactivity worth of its control rods in order to predict the behavior of the reactor. Westinghouse Savannah River Company (WSRC) uses the GLASS lattice-physics code to calculate few-group cross sections for fuel and control-rod assemblies in the K reactor. This paper compares the control-rod worth calculated by GLASS to that calculated by the MCNP Monte Carlo program. The GLASS calculations utilize its standard 37-group cross-section library, while the MCNP calculations employ continuous-energy isotopic cross-section libraries derived from ENDF/B-V. The MCNP calculations therefore combine the most rigorous analytical model and the most accurate cross sections currently available for thermal-reactor analysis. Consequently, the MCNP results comprise a computational benchmark against which the accuracy of the GLASS code can be evaluated

  10. Study on dosimetric properties of radiophotoluminescent glass rod detector

    Energy Technology Data Exchange (ETDEWEB)

    Rah, Jeong Eun; Hong, Ju Young; Suh, Tea Suk [The Catholic University of Korea, Seoul (Korea, Republic of); Shin, Dong Oh [Kyunghee Univ., Seoul (Korea, Republic of); Kim, Hee Sun [Korea Hydro and Nuclear Power Co., Ltd, Daejeon (Korea, Republic of); Lim, Chun Il; Jeong, Hee Gyo [Korea Food and Drug Administration, Seoul (Korea, Republic of)

    2006-12-15

    A radiophotoluminescent Glass Rod Detector (GRD) system has recently become commercially available. We investigate the dosimetric properties of the GRD regarding the reproducibility of signal, dose linearity and energy dependence. The reproducibility of five measurements for 50 GRDs is presented by an average of one standard deviation of each GRD and it is {+-}1.2%. It is found to be linear in response to doses of {sup 60}Co beam in the range 0.5 to 50 Gy with a coefficient of linearity of 0.9998. The energy dependence of the GRD is determined by comparing the dose obtained using cylindrical chamber to that by using the GRD. The GRD response for each beam is normalized to the response for a {sup 60}Co beam. The responses for 6 and 15 MV x-ray beams are within {+-}1.5% (1SD). The energy response of GRD for high-energy photon is almost the same as the energy dependence of LiF:Mg:Ti (TLD-100) and shows little energy dependence unlike p-type silicon diode detector. The GRDs have advantages over other detectors such diode detector, and TLD: linearity, reproducibility and energy dependency. It has been verified to be an effective device for small field dosimetry for stereotactic radiosurgery.

  11. The influence of the energy distribution of workplace fields on neutron personal dosemeter reading

    International Nuclear Information System (INIS)

    Tanner, R.J.; Thomas, D.J.; Bartlett, D.T.; Hager, L.G.; Horwood, N.A.

    2002-01-01

    Variations in the energy dependence of response of neutron personal dosemeters cause systematic errors in the readings obtained in workplace fields. The magnitude of these errors has been determined theoretically by folding measured and calculated workplace energy distributions with dosemeter response functions, to determine the response of a given personal dosemeter in that field. These results have been analysed with consideration of the dosemeter response to various calibration spectra, and with reference to different workplaces. The dosemeters in the study are discussed in terms of the workplaces for which they can be suitably calibrated. Deficiencies in the published neutron energy distributions are identified

  12. Dosimetric and kinetic parameters of lithium cadmium borate glasses doped with rare earth ions

    Directory of Open Access Journals (Sweden)

    J. Anjaiah

    2014-10-01

    Full Text Available Thermoluminescence (TL characteristics of X-ray irradiated pure and doped with four different rare earth ions (viz., Pr3+, Nd3+, Sm3+ and Eu3+ Li2O–Cdo–B2O3 glasses have been studied in the temperature range 303–573 K; the pure glass has exhibited single TL peak at 466 K. When this glass is doped with different rare earth ions no additional peaks are observed but the glow peak temperature of the existing glow peak shifted gradually towards higher temperatures with gain in intensity of TL light output. The area under the glow curve is found to be maximum for Eu3+ doped glasses. The trap depth parameters associated with the observed TL peaks have been evaluated using Chen's formulae. The possible use of these glasses in radiation dosimetry has been described. The result clearly showed that europium doped cadmium borate glass has a potential to be considered as the thermoluminescence dosimeter.

  13. Measurements of radon in dwellings with CR-39 track detectors

    DEFF Research Database (Denmark)

    Majborn, Benny

    1986-01-01

    A passive integrating dosemeter has been designed for measuring natural radiation in dwellings. The dosemeter contains one or two CR-39 track detectors to measure radon and three thermoluminescence dosemeters to measure external radiation. The dosemeter was investigated in a pilot study in 1983....../84, and it is now used in a nationwide survey of natural radiation in Danish dwellings. The characteristics of the dosemeter with respect to radon measurements are presented, and the radon monitoring results obtained in the pilot study are summarized...

  14. Fracture detection in concrete by glass fiber cloth reinforced plastics

    Science.gov (United States)

    Shin, Soon-Gi; Lee, Sung-Riong

    2006-04-01

    Two types of carbon (carbon fiber and carbon powder) and a glass cloth were used as conductive phases and a reinforcing fiber, respectively, in polymer rods. The carbon powder was used for fabricating electrically conductive carbon powder-glass fiber reinforced plastic (CP-GFRP) rods. The carbon fiber tows and the CP-GFRP rods were adhered to mortar specimens using epoxy resin and glass fiber cloth. On bending, the electrical resistance of the carbon fiber tow attached to the mortar specimen increased greatly after crack generation, and that of the CP-GFRP rod increased after the early stages of deflection in the mortar. Therefore, the CP-GFRP rod is superior to the carbon fiber tow in detecting fractures. Also, by reinforcing with a glass fiber cloth reinforced plastic, the strength of the mortar specimens became more than twice as strong as that of the unreinforced mortar.

  15. A study of the irradiation temperature coefficient for L-alanine and DL-alanine dosemeters

    International Nuclear Information System (INIS)

    Desrosiers, M. F.; Lin, M.; Cooper, S. L.; Cui, Y.; Chen, K.

    2006-01-01

    Alanine dosimetry is now well established both as a reference and routine dosemeter for industrial irradiation processing. Accurate dosimetry under the relatively harsh conditions of industrial processing requires a characterisation of the parameters that influence the dosemeter response. The temperature of the dosemeter during irradiation is a difficult quantity to measure so that the accuracy of the temperature coefficient that governs the dosemeter response becomes a critical factor. Numerous publications have reported temperature coefficients for several types of alanine dosemeters. The observed differences in the measured values were commonly attributed to the differences in the polymer binder or the experimental design of the measurement. However, the data demonstrated a consistent difference in the temperature coefficients between L-alanine and DL-alanine. Since there were no commonalities in the dosemeter composition or the measurement methods applied, a clear conclusion is not possible. To resolve this issue, the two isomeric forms of alanine dosemeters were prepared and irradiated in an identical manner. The results indicated that the DL-alanine temperature coefficient is more than 50% higher than the L-alanine temperature coefficient. (authors)

  16. Investigation about the thermal features of the ovens used for thermoluminescence; Indagine sulle specifiche dei forni usati per i trattamenti termici dei rilevatori a termoluminescenza

    Energy Technology Data Exchange (ETDEWEB)

    Scarpa, G; Caporali, C; Moscati, M [ENEA - Area Energia, Ambiente e Salute, Centro Ricerche Energia, Casaccia, Rome (Italy)

    1991-02-15

    The present paper reports the results of an investigation carried out by the PAS-FIBI-DOSIBIO laboratory (ENEA, Casaccia, Roma) about the thermal features of the ovens used for annealing treatments of TL dosemeters. A total number of 45 commercial ovens and muffle furnaces were studied. belonging to 24 Italian Health Physics laboratories. The investigation has shown that the majority of the ovens do not possess a degree of accuracy, stability, uniformity and reproducibility suitable for their use in the field of thermoluminescence dosimetry. Practical suggestions are also given in order to reduce the effects of some of the negative characteristics found in most ovens. (author)

  17. Personal radiation monitoring

    International Nuclear Information System (INIS)

    Julius, H.W.

    1976-01-01

    This report reviews the context of a primary university course in individual radiation monitoring. A brief account of the regulations and permissible doses is given. The principles and design of film dosemeters, thermoluminescent dosemeters and the whole-body counting technique are treated

  18. An investigation into the effect of protective devices on the dose to radiosensitive organs in the head and neck

    International Nuclear Information System (INIS)

    Marshall, N.W.; Faulkner, K.; Clarke, P.

    1992-01-01

    A series of experiments were performed to determine the dose reduction afforded to radiosensitive organs in the head and neck by various protective devices. These included spectacles with plastic, standard glass, photochromic and lead-glass lenses, a thyroid collar and a lead-acrylic face mask. The measurements were performed using an anthropomorphic phantom loaded with lithium fluoride thermoluminescent dosemeters, in conditions realistic of clinical practice. Irradiations were performed using scattered radiation produced by a pelvic phantom, for X-ray beams generated at 80 kVp and 110 KVp. Also presented is the ratio of organ dose to dose to the bridge of the nose for thyroid, oesophagus, brain and sinuses, as measured for the case of no head or neck protection. (author)

  19. Thermoluminescent dosimetric properties of Descalvado sand

    International Nuclear Information System (INIS)

    Teixeira, M.I.; Caldas, L.V.E.

    2006-01-01

    Sand samples proceeding from Descalvado, Sao Paulo, were studied with regard to their dosimetric properties using the thermoluminescence technique (TL) for high doses. These sand samples present steady physical and chemical characteristics to the end items, and they are used in the glass industry and for casting. The TL curves of the samples were obtained after an irradiation at the Gamma-Cell system ( 60 Co), of IPEN. The glow curves present two peaks at 80 C and 220 C approximately. Calibration curves were obtained for doses between 50 Gy and 5 kGy. The results indicate that the sand samples can be used for high-doses dosimetry in several areas of applications of ionizing radiation. (Author)

  20. Thermoluminescent dosimetric properties of Descalvado sand

    Energy Technology Data Exchange (ETDEWEB)

    Teixeira, M.I.; Caldas, L.V.E

    2006-07-01

    Sand samples proceeding from Descalvado, Sao Paulo, were studied with regard to their dosimetric properties using the thermoluminescence technique (TL) for high doses. These sand samples present steady physical and chemical characteristics to the end items, and they are used in the glass industry and for casting. The TL curves of the samples were obtained after an irradiation at the Gamma-Cell system ({sup 60} Co), of IPEN. The glow curves present two peaks at 80 C and 220 C approximately. Calibration curves were obtained for doses between 50 Gy and 5 kGy. The results indicate that the sand samples can be used for high-doses dosimetry in several areas of applications of ionizing radiation. (Author)

  1. Effect of Dy{sub 2}O{sub 3} impurities on the physical, optical and thermoluminescence properties of lithium borate glass

    Energy Technology Data Exchange (ETDEWEB)

    Mhareb, M.H.A., E-mail: mmhareb@hotmail.com [Radiation Protection Directorate, Energy and Minerals Regulatory Commission, Amman 11821 (Jordan); Hashim, S. [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 UTM Skudai, Johor (Malaysia); Ghoshal, S.K., E-mail: lordshib@gmail.com [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 UTM Skudai, Johor (Malaysia); Alajerami, Y.S.M. [Department of Medical Radiography, Al-Azhar University, Gaza Strip, Palestine (Country Unknown); Bqoor, M.J.; Hamdan, A.I. [Radiation Protection Directorate, Energy and Minerals Regulatory Commission, Amman 11821 (Jordan); Saleh, M.A. [Nuclear Engineering Programme, Faculty of Petroleum and Renewable Energy Engineering, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Karim, M.K.B. Abdul [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 UTM Skudai, Johor (Malaysia)

    2016-09-15

    Dysprosium (Dy) doped lithium borate glass (LBG) is prepared using conventional melting-quenching technique with varying Dy concentration in the range of 0 to 1.0 mol%. Prepared glass samples are characterized via X-Ray Diffraction (XRD), Fourier Transform Infrared (FTIR), UV–vis–IR, Photoluminescence (PL), Thermoluminescence (TL) spectroscopy and Differential Thermal Analysis (DTA). The physical parameters such as the density, optical energy band gap, oscillator strength, refractive index, ion concentration, Polaron radius, molar volume and inter-nuclear distance are determined. UV–vis–IR spectra revealed seven prominent bands centered at 448, 749, 796, 899, 1085, 1265 and 1679 nm corresponding to the transition from the Dy{sup 3+} ion's ground state ({sup 6}H{sub 15/2}) to the excited states ({sup 4}I{sub 15/2},{sup 6}F{sub 3/2}, {sup 6}F{sub 5/2}, {sup 6}H{sub 5/2}, {sup 6}F{sub 9/2}, {sup 6}H{sub 9/2} and {sup 6}H{sub 11/2}). The room temperature photoluminescence (PL) spectra of the glass series at 350 nm excitation displayed two peaks centered at 481 nm and 573 nm, which are assigned to the transitions of ({sup 4}F{sub 9/2}→{sup 6}H{sub 15/2}) and ({sup 4}F{sub 9/2}→{sup 6}H{sub 13/2}), respectively. The TL spectra of gamma-irradiated samples are measured, which showed Dy{sup 3+} content dependent simple glow peak at 190 °C. Dysprosium ion is found to play an important role in the TL and PL intensity enhancement of LB phosphor.

  2. Glass forming ability and mechanical properties of Zr50Cu42Al8 bulk metallic glass

    International Nuclear Information System (INIS)

    Xia, L; Chan, K C; Wang, G; Liu, L

    2008-01-01

    In this work, we report that Zr 50 Cu 42 Al 8 bulk metallic glass (BMG) exhibits excellent glass forming ability and mechanical properties. Zr 50 Cu 42 Al 8 glassy rods with a diameter of 3 mm were prepared using conventional copper mould suction casting. The glassy rod exhibits a modulus of about 115 GPa and a fracture strength of about 2 GPa, and, as compared with other large-scale BMGs, it has excellent room-temperature plasticity of up to 20% under compression. The fracture mechanism of the rod was investigated by microstructural investigations, and it was found that the large plasticity of the as-cast rod is closely related to the in situ formation of nano-crystalline particles embedded in the amorphous matrix.

  3. Thermoluminescent dosimetry in dentistry students in radiological training; Dosimetria termo-luminiscente en estudiantes de odontologia en entrenamiento radiologico

    Energy Technology Data Exchange (ETDEWEB)

    Loya, M.; Duarte, R. [Escuela de Odontologia, Universidad Autonoma de Chihuahua (Mexico); Montero, M.E. [CIMAV, Chihuahua (Mexico); Gonzalez, P.R. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico); Ojeda, S.L.; Sanin, L.H. [Facultad de Enfermeria y Nutriologia, Chihuahua (Mexico)

    2007-07-01

    Full text: In this work, the obtained results of the mensuration of the equivalent one are presented of dose (E), in students of the career of Dentistry who during their training in radiology complete the three lists of patient, odontologist and observer. The study one carries out with 35 students (Odontologists), 7 men and 28 women, with ages between 20 and 24 years. A characterization of risk was made following the methodology of the Agency for the environmental control of United States (EPA), counting the number of shots in each list, during the time of training. The effective energy of the X rays used was of 24 KeV. The irradiation technique understood two modalities; the first one with 3 shots of 1.25 seconds and the second with 5 shots of 0.6 seconds. The measurement of E, was carried out with thermoluminescent dosemeters (TL) of LiF:Mg,Cu,P+PTFE, developed in the National Institute of Nuclear Research (ININ). The homogenization of the dosemeters showed a variation in its response to the radiation of 1.85%DS. For the odontologist position, the dosemeters was placed to the height of the chest and the measurement was multiplied by the number of shots that it was 304. The E value was of 3.62 mSv/four-month (11 mSv/year), without any safety equipment and of 2.02 mSv/four-month (6.14 mSv/year), when was considered the scenario for this position with use of vest and lead collar, like safety equipment, diminishing E, almost 50%. In the case of the corresponding position to patient, E, it was measured in different organs of interest, the obtained results are compared with the international standards. Recommendations are given for similar cases. (Author)

  4. A gamma/neutron-discriminating, Cooled, Optically Stimulated Luminescence (COSL) dosemeter

    International Nuclear Information System (INIS)

    Eschbach, P.A.; Miller, S.D.

    1992-07-01

    The Cooled Optically Stimulated Luminescence (COSL) of CaF 2 :Mn (grain sizes from 0.1 to 100 microns) powder embedded in a hydrogenous matrix is reported as a function of fast-neutron dose. When all the CaF 2 :Mn grains are interrogated at once, the COSL plastic dosemeters have a minimum detectable limit of 1 cSv fast neutrons; the gamma component from the bare 252 cf exposure was determined with a separate dosemeter. We report here on a proton-recoil-based dosemeter that generates pulse height spectra, much like the scintillator of Hornyak, (2) to provide information on both the neutron and gamma dose

  5. Numerical and experimental results of the operational neutron dosemeter 'Saphydose-N'

    International Nuclear Information System (INIS)

    Lahaye, T.; Chau, Q.; Menard, S.; Ndontcheung-Moyo, M.; Bolognese-Milsztajn, T.; Rannou, A.

    2004-01-01

    Since 1993, the Inst. for Radiological Protection and Nuclear Safety (IRSN) has lead, in association with Electricite de France (EDF), a R and D study of a neutron personal electronic dosemeter. This dosemeter, called 'Saphydose-N', is manufactured by the SAPHYMO company. This paper presents first the optimisation of some detector components using Monte Carlo calculations, and second the test of the manufactured Saphydose-N under radiation following the IEC 1323 standard's recommendations for active personal neutron dosemeters. The measurements with the manufactured dosemeter were performed on the one hand at PTB (Physikalisch-Technische Bundesanstalt) in mono-energetic neutron fields and, on the other hand at IRSN in neutron fields generated by a thermal facility (SIGMA), radionuclide ISO sources and a realistic spectrum (CANEL/T400). The manufactured dosemeter Saphydose-N was also tested during measurement campaigns of the European programme EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields') at different nuclear workplaces. The study showed that Saphydose-N complies with the recommendations of standard IEC 1323 and can be used at any workplace with no previous knowledge of the neutron field characteristics. (authors)

  6. Proposal of a dosemeter for skin beta radiation dose assessment

    International Nuclear Information System (INIS)

    Rosa, L.A.R. da; Caldas, L.V.E.

    1987-08-01

    Beta radiation is, undoubtedly, less penetrating than X or gamma radiation. Thus, beta radiation sources external to the human body do not cause a significant irradiation of its deeper tissues. However, in some cases, they may contribute in a very important way to the irradiation of the lens of the eyes and, mainly, of the skin. Specially, the hands and finger tips may receive a high dose. In this work some relevant aspects of the individual monitoring in beta radiation fields are discussed and the importance of monitoring this kind of radiation in some activities where the skin absorbed dose may be a limiting factor is evidenced. The main characteristics of the thermoluminescent (TL) response of ultra-thin CaSO 4 : Dy detectors (UT-CaSO 4 : Dy) in the detection of this kind of radiation are also studied. The irradiation are performed with 90 Sr 90 Y, 204 TI and 147 Pm sources. The reproducibility, linearity, dependence on the absorbed dose rate, optical fading, energy and angular dependences of the detector TL responce are investigated. Transmission factors for different thicknesses of tissue equivalent material are obtained for the TL detectors using the three available beta sources. Based on the results obtained, a dosemeter for skin beta radiation absorbed dose assessment with an energy dependence better than 12% is proposed. (Author) [pt

  7. Dosimetric properties of dysprosium doped lithium borate glass irradiated by 6 MV photons

    International Nuclear Information System (INIS)

    Ab Rasid, A.; Wagiran, H.; Hashim, S.; Ibrahim, Z.; Ali, H.

    2015-01-01

    Undoped and dysprosium doped lithium borate glass system with empirical formula (70–x) B 2 O 3 –30 Li 2 O–(x) Dy 2 O 3 (x=0.1, 0.3, 0.5, 0.7, 1.0 mol%) were prepared using the melt-quenching technique. The dosimetric measurements were performed by irradiating the samples to 6 MV photon beam using linear accelerator (LINAC) over a dose range of 0.5–5.0 Gy. The glass series of dysprosium doped lithium borate glass produced the best thermoluminescence (TL) glow curve with the highest intensity peak from sample with 1.0 mol% Dy 2 O 3 concentration. Minimum detectable dose was detected at 2.24 mGy, good linearity of regression coefficient, high reproducibility and high sensitivity compared to the undoped glass are from 1.0 mol% dysprosium doped lithium borate glass. The results indicated that the series of dysprosium doped lithium glasses have a great potential to be considered as a thermoluminescence dosimetry (TLD). - Highlights: • TL response of undoped and dysprosium doped lithium borate glass subjected to 6 MV photons irradiation at low dose range. • TL linear response of dysprosium doped lithium borate glass. • The sensitivity of dysprosium doped lithium borate glass is approximately 93 times higher than undoped glass

  8. Thermoluminescence, luminescence optically stimulated and creation of defects in alkaline halogenides contaminated with Europium

    International Nuclear Information System (INIS)

    Barboza F, M.

    1999-01-01

    The alkaline halogenides have been subject matter of investigations related with the search of sensor materials for X-ray bidimensional images or optical memories. The understanding of the damage formation processes generated by ionizing and non-ionizing radiations is important for the correct design of devices that working as detectors and dosemeters of both type of radiations. In this work we present the investigation results related with the defects produced by the ionizing radiation type X and ultraviolet light in the range of 200-360 nm in crystals of KCl: Eu 2+ and KBr:Eu 2+ . It is examined the thermoluminescence and luminescence spectra with the purpose of identifying the exciton processes, owing to the excitation of the halogenide ions in which the primary defects correspond to the F and H centers. It has been found that the 400-600 nm emission is associated with the luminescence type that in his turn can be associated with autotrapped excitons perturbed by the impurity. On the other hand, it is examined the emission spectra of the luminescence optically stimulated in crystals of KBr: Eu 2+ and KCl: Eu 2+ finding too that such materials would be used as optical memories susceptible of to store information, and through of photostimulation to read this. It was determined that the F centers participate in the luminescence optically stimulated in these crystals, as well as too in the recombination processes responsible by the thermoluminescent emission. (Author)

  9. Thermoluminescence and optically stimulated luminescence properties of Dy{sup 3+}-doped CaO–Al{sub 2}O{sub 3}–B{sub 2}O{sub 3}-based glasses

    Energy Technology Data Exchange (ETDEWEB)

    Yahaba, T., E-mail: takuma.yahaba.s1@dc.tohoku.ac.jp [Department of Applied Chemistry, Graduate School of Engineering, Tohoku University, 6-6-07 Aramaki Aza Aoba, Aoba-ku, Sendai 980-8579 (Japan); Fujimoto, Y. [Department of Applied Chemistry, Graduate School of Engineering, Tohoku University, 6-6-07 Aramaki Aza Aoba, Aoba-ku, Sendai 980-8579 (Japan); Yanagida, T. [Nara Institute of Science and Technology (NAIST), 8916-5 Takayama, Ikoma 630-0192 (Japan); Koshimizu, M.; Tanaka, H.; Saeki, K.; Asai, K. [Department of Applied Chemistry, Graduate School of Engineering, Tohoku University, 6-6-07 Aramaki Aza Aoba, Aoba-ku, Sendai 980-8579 (Japan)

    2017-02-01

    We developed Dy{sup 3+}-doped CaO–Al{sub 2}O{sub 3}–B{sub 2}O{sub 3} based glasses with Dy concentrations of 0.5, 1.0, and 2.0 mol% using a melt-quenching technique. The as-synthesized glasses were applicable as materials exhibiting thermoluminescence (TL) and optically stimulated luminescence (OSL). The optical and radiation response properties of the glasses were characterized. In the photoluminescence (PL) spectra, two emission bands due to the {sup 4}F{sub 9/2} → {sup 6}H{sub 15/2} and {sup 4}F{sub 9/2} → {sup 6}H{sub 13/2} transitions of Dy{sup 3+} were observed at 480 and 580 nm. In the OSL spectra, the emission band due to the {sup 4}F{sub 9/2} → {sup 6}H{sub 15/2} transition of Dy{sup 3+} was observed. Excellent TL and OSL responses were observed for dose ranges of 0.1–90 Gy. In addition, TL fading behavior was better than that of OSL in term of the long-time storage. These results indicate that the Dy{sup 3+}-doped CaO–Al{sub 2}O{sub 3}–B{sub 2}O{sub 3}-based glasses are applicable as TL materials.

  10. The right choice: extremity dosemeter for different radiation fields

    International Nuclear Information System (INIS)

    Brasik, N.; Stadtmann, H.; Kindl, P.

    2005-01-01

    Full text: Measurements of weakly penetrating radiation in personal dosimetry present problems in the design of suitable detectors and in the interpretation of their readings. For the measurement of the individual beta radiation dose, personal dosemeter for the fingers/tips are required. In general, the dosemeters currently used for personal monitoring of beta and low energy photon doses suffer from an energy threshold problem because the detector and/or the filter are too thick. TLDs of a standard thickness can seriously underestimate personal skin doses, especially in external fields of weakly penetrating radiation fields. LiF:Mg, Cu, P is a promising TL material which allows the production of thin detectors with sufficient sensitivity. Dosimetric properties of two different types of extremity dosemeters, designed to measure the personal dose equivalent Hp(0.07), have been compared: LiF:Mg, Ti (TLD100) and LiF:Mg, Cu, P (TLD700H). The first one consists of 100 mg.cm -2 LiF:Mg, Ti (TLD 100) chip and a 35mg. cm -2 cap, the other consists of a 7mg. cm -2 layer of LiF:Mg, Cu, P (TLD-700H) powder and a 5mg. cm -2 cap. The evaluation was done in two steps: performance tests (ISO-12794) and measurements in real workplaces. In the first step type test results for beta calibration were compared. In addition calibration for low energy photon radiation according to ISO 4037-3 was carried out. In the second step, simultaneous measurements with both types of dosemeters were performed at workplaces, where radiopharmaceuticals containing different radioisotopes are prepared and applied. Practices in these fields are characterized by handling of high activities at very small distances between source and skin. The results from the comparison of the two dosemeter types are presented and analyzed with respect to different radiation fields. Experiments showed a satisfactory sensitivity for the thinner dosemeter (TLD 700H) for detecting beta radiation at protection levels and a good

  11. The NRPB nuclear emission dosemeter

    International Nuclear Information System (INIS)

    Bartlett, D.T.; Bird, T.V.; Miles, J.C.H.

    1980-03-01

    The NRPB nuclear emulsion dosemeter is designed such that the thermal neutron response is similar to the fast neutron response. This report gives details of the dependence of response on incident neutron energy and on neutron angle of incidence. (author)

  12. Evaluation of personal dosemeters and comparative measurements performed by PTB in the years 1979 through 1984. Pt. 1

    International Nuclear Information System (INIS)

    Boehm, J.; Buchholz, G.

    1986-01-01

    Comparison measurements of the official individual dosemeters performed in the PTB from 1979 to 1984 are described. Also a number of other, unofficial individual dosemeters participated. The performance of the comparison measurements and the types of the individual dosemeters are described. Tables show the values measured with the film dosemeters, with film and TLD dosemeters, and TLD dosemeters as well as dosimetric probes for whole-body and partial body dose measurement. (DG) [de

  13. THERMOLUMINESCENT DOSIMETRY

    Energy Technology Data Exchange (ETDEWEB)

    Malsky, Stanley J.; Amato, Charles G.; Roswit, Bernard

    1963-06-15

    The performance of CaF/sub 2/: Mn and LiF as thermoluminescent in vivo dosimeters is studied. The facility with which this type of dosimeter can be inserted into the body is discussed. The dosimeter is inserted into the body by trocar or catheter techniques and exposed to radiation; it is then removed and heated, and the thermoluminescence is measured and used as an indication of the absorbed dose. (T.F.H.)

  14. Leaked water detection device for control rod drive and BWR type reactor

    International Nuclear Information System (INIS)

    Takahashi, Ken.

    1995-01-01

    The device of the present invention can specify a control rod drive causing great amount of water leakage among a large number of control rod drives. Namely, water leaked from the control rod drives is introduced to each of leaked water pipelines. Further, it is introduced from the leaked water pipelines to flow glasses at which leaked water can visually be recognized individually, and then discharged through a drain pipeline. With such procedures, the amount of leaked water from the leaked water pipelines can visually be recognized at the flow glasses. As a result, the control rod drives which cause a great amount of leakage can be specified among large number of control rod drives. Accordingly, an accurate inspection schedule for a shaft-sealing portion of the control rod drives can be formed. The shaft-sealing portion degradated in the sealing property can reliably be inspected and repaired. Purge water can be ensured to improve reliability of the operation of equipments. (I.S.)

  15. Thermoluminescent dosimetry: theory, materials and applications

    International Nuclear Information System (INIS)

    Rosa, L.A.R.da.

    1978-01-01

    A survey about the thermoluminescence theory, the properties of the main thermoluminescent phosphors and their applications is presented. Some of the most important thermoluminescent readers are also mentioned [pt

  16. Application of chemical dosemeters during the irradiation of agricultural products

    Energy Technology Data Exchange (ETDEWEB)

    Kovacs, A; Stenger, V; Foeldiak, G [Magyar Tudomanyos Akademia, Budapest. Izotopintezete

    1983-09-01

    The joint recommedation by the FAO, WHO and IAEA in 1981 proposed the unrestricted permission of food irradiation under the dose value of 10 kGy (1 Mrd). The chemical dosemeters presented in the paper cover the dose range of 30 Gy to 1 MGy for use to food irradiation safety measurements. The Fricke dosemeter was used in the 50-400 Gy range at radappertization operations of onion, poppy and potato. The alcoholic chlorobenzene dosemeter with oscillometric analysis was applied successfully at low (0.4-4 kGy) ranges, primarily for irradiation dose monitoring of fruits, vegetables, refrigerated poultry etc. The conductometry could be used for dosimetry in the range of 60 Gy to 1 MGy.

  17. Analysis of environmental impact due to gaseous wastes releasing during Angra-1 operation

    International Nuclear Information System (INIS)

    Cruz, E.S. da; Kircher, E.

    1986-01-01

    The results of environmental monitoring, carried out in six terrestrial stations around the Angra-1 reactor are presented. The radiation doses were measured using calcium fluoride thermoluminescent dosemeters. A statistical treatment for data obtained by dosemeters to avoid background radiations was done. (M.C.R.) [pt

  18. Glow curve characteristics of bulb type thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Deme, S.; Feher, I.; Felszerfalvi, J.

    1993-01-01

    TL dosemeter readers are equipped usually with thermocouples connected to the heater unit. This layout can well be applied to stabilize the position of the glow curve as a function of heating-up time. Bulb type TL dosemeters do not have temperature sensors, no possibility for stabilization, which can cause an additional readout error of dose determination. For this reason, the time dependence of glow curves for bulb-type TL dosemeters was measured, and a new microprocessor controlled readout device was developed. (N.T.) 2 refs.; 2 figs

  19. Comparative study of thermoluminescent properties of LiF: Mg, Cu, P, LiF: Mg, Ti and TLD-100 irradiated with X-rays

    International Nuclear Information System (INIS)

    Azorin, J.; Rivera, T.; Gonzalez, P.; Ortega, X.; Ginjaume, M.

    2000-01-01

    The thermoluminescent properties (Tl) of LiF: Mg, Cu, P, and LiF: Mg, Ti, were investigated both developed in Mexico and comparing them with the properties of TLD-100 when they are exposure to X-rays. The Tl curve of LiF: Mg, Cu, P exhibited two peaks at 200 and 300 Centigrade. Its response Tl in function of dose resulted linear in the interval of 0.5 Gy until 5 Gy and its sensitivity to X-ray was around 25 times greater that of the TLD-100. Also it was measured the Tl response of the three materials in function of photon energy. The results showed that LiF: Mg, Cu, P has potential to be used as X-ray dosemeter. (Author)

  20. Radiation effects in the polycarbonate of bisphenol-A. Thermoluminescence electron spin resonance and charged particle track studies

    International Nuclear Information System (INIS)

    Edmonds, E.A.

    1978-09-01

    A detailed investigation is presented of the thermoluminescence observable above room temperature from the polycarbonate of bisphenol-A after its exposure to different radiations. A correlation study is described by which features of the complex thermoluminescence glow curve from a commercial grade of the polycarbonate of bisphenol-A are related to the etchability of charged particle damage trails and the radiogenic ESR signal. A model is presented whereby the etchability of charged particle damage trails is associated with chain scission caused by the high local dose of radiation in the vicinity of the trajectories of charged particles. Methods by which activation constants controlling the thermoluminescence glow curve can be evaluated are discussed and results are presented. It is concluded that glow peaks associated with the ESR signal or enhanced etchability are related to small-scale motions in the molecular matrix of the polycarbonate of bisphenol-A. These motions are thermally activated in accord with the simple Boltzmann relation usually incorporated into theories of thermoluminescence. Another component glow peak of the thermoluminescence glow curve is shown to be associated with the glass-rubber transition in the polycarbonate of bisphenol-A. Different features of the thermoluminescence glow curve can be related to relaxations of the polymer matrix and decomposition of the matrix. It is confirmed that the dominant bulk effect of radiation in the polycarbonate of bisphenol-A exposed to large doses of radiation is chain scission. (author)

  1. Thermoluminescence dating method

    International Nuclear Information System (INIS)

    Zink, A.

    2004-01-01

    A crystal that is submitted to radiation stores energy and releases this energy under the form of light whenever it is heated. These 2 properties: the ability to store energy and the ability to reset the energy stored are the pillars on which time dating methods like thermoluminescence are based. A typical accuracy of the thermoluminescence method is between 5 to 7% but an accuracy of 3% can be reached with a sufficient number of measurement. This article describes the application of thermoluminescence to the dating of a series of old terra-cotta statues. This time measurement is absolute and does not require any calibration, it represents the time elapsed since the last heating of the artifact. (A.C.)

  2. Beam depolarization and gain saturation in neodymium rods with a diameter of 85 mm

    Energy Technology Data Exchange (ETDEWEB)

    Sukhanov, V N; Ugodenko, A A

    1990-04-01

    Depolarization and gain saturation were investigated in rod amplifiers using phosphate and silicate Nd glasses 85 mm in diameter and 300 mm in length at a pulse duration of 35 ns. Total depolarization losses over the rod cross section were measured for various radial distributions of the small-signal gain. For the phosphate glass the losses amounted to 3-6 percent; for the silicate glass, they amounted to 4-7 percent. Saturation energy densities of 4.5 + or - 0.4 and 8.0 + or - 0.7 J/sq cm were obtained for the phosphate and silicate glass, respectively. 8 refs.

  3. EURADOS intercomparison 2012 for neutron dosemeters

    International Nuclear Information System (INIS)

    Fantuzzi, E.; Chevallier, M.A.; Cruz-Suarez, R.; Luszik-Bhadra, M.; Mayer, S.; Thomas, D.J.; Tanner, R.; Vanhavere, F.

    2014-11-01

    EURADOS, within the work performed by Working Group 2 - Harmonization of Individual Monitoring in Europe, has started a self-sustained programme of regular intercomparisons and has successfully executed three intercomparisons for whole body photon dosemeters (IC2008, IC2010, IC2012) and one intercomparison for extremity dosemeters for photon and beta fields (IC2009). In 2012, the EURADOS intercomparison IC2012n was launched for personal neutron dosemeters routinely used to measure personal dose equivalent, Hp(10), in individual monitoring. No systems under development were allowed to participate. IC2012n was carried out by a EURADOS nominated Organization Group (OG) consisting of: Marie- Anne Chevallier (IRSN, F), Rodolfo Cruz-Suarez (IAEA, UN-Vienna), Marlies Luszik-Bhadra (PTB, D), Sabine Mayer (PSI, CH), David J. Thomas (NPL, UK), Rick Tanner (PHE, UK), Filip Vanhavere (SCK-CEN, B) led by a Coordinator, Elena Fantuzzi (ENEA, I). 31 participants registered for the comparison, with 34 dosimetry systems. In total 816 dosemeters were irradiated in selected neutron fields on an ISO slab phantom. The irradiations were performed at 2 European accredited laboratories which are both National Primary Metrology Laboratories for ionizing radiation: NPL (National Physical Laboratory, UK) and PTB (Physikalisch-Technische Bundesanstalt, D). All irradiations were carried out according to the irradiation plan developed by the OG. The overall results show that most, although not all, dosimetric systems perform acceptably well (within a factor of 2) for irradiations with a bare radionuclide source ( 252 Cf at 0 ), whilst more than half of the systems underestimate the reference value for irradiations from non-normal angles of incidence irradiations ( 252 Cf at 45 ) or for simulated workplace fields ( 252 Cf(D 2 O) or 252 Cf source behind a shadow cone). The performance for 250 keV mono-energetic neutron irradiations varies mainly reflecting the detection principle on which the

  4. Thermoluminescence in CVD diamond films: application to actinometric dosimetry

    International Nuclear Information System (INIS)

    Barboza-Flores, M.; Melendrez, R.; Chernov, V.; Castaneda, B.; Pedroza-Montero, M.; Gan, B.; Ahn, J.; Zhang, Q.; Yoon, S.F.

    2002-01-01

    Diamond is considered a tissue-equivalent material since its atomic number (Z=6) is close to the effective atomic number of biological tissue (Z=7.42). Such a situation makes it suitable for radiation detection purposes in medical applications. In the present work the analysis is reported of the thermoluminescence (TL) and dosimetric features of chemically vapour deposited (CVD) diamond film samples subjected to ultraviolet (UV) irradiation in the actinometric region. The TL glow curve shows peaks at 120, 220, 320 and 370 deg. C. The 120 and 370 deg. C peaks are too weak and the first one fades away in a few seconds after exposure. The overall room temperature fading shows a 50% TL decay 30 min after exposure. The 320 deg. C glow peak is considered to be the most adequate for dosimetric applications due to its low fading and linear TL behaviour as a function of UV dose in the 180-260 nm range. The TL excitation spectrum presents a broad band with at least two overlapped components around 205 and 220 nm. The results indicate that the TL behaviour of CVD diamond film can be a good alternative to the currently available dosemeter and detector in the actinometric region as well as in clinical and medical applications. (author)

  5. Thermoluminescent dosimeter system

    International Nuclear Information System (INIS)

    Felice, P.E.; Gonzalez, J.L.; Seidel, J.G.

    1979-01-01

    An improved thermoluminescent dosimeter system and apparatus for sensing alpha particle emission is described. A thermoluminescent body is sealed between a pair of metallized plastic films. The dosimeter is mounted within a protective inverted cup or a tube closed at one end, which is disposed in a test hole for exposure to radioactive radon gas which is indicaive of uranium deposits

  6. Radiological environmental monitoring with LiF-700-H dosemeters

    International Nuclear Information System (INIS)

    Scarnichia, E.; Levanon, I.; Andres, P.; Miani, C.; Ramirez, S.

    2011-10-01

    Since 2008 a radiological environmental monitoring with LiF-700-H has been carried out as a result of increasing the Ra-6 research reactor core power. The information obtained is used to evaluate and to quantify analytically the air kerma rate, the fading and the associated uncertainty by developing software tools (deconvolution and uncertainty algorithms). LiF-700-H dosemeters were chosen because of their high sensitivity to low air kerma rates. They show a very good stability and a negligible fading for two-month working periods. The air kerma rate detection limit (based on the 3σ criterion) during these working periods is about 0.4 n Gy/h. Air kerma rates of about 70 n Gy/h are measured with this detection limit. Following the Nist guidelines, an algorithm was developed in order to find the associated uncertainty. It considers several aspects, such as the source activity decay, distance source-dosemeter during the calibration procedure, irradiation time, calibration factor, dosemeter readout, dosemeter sensitivity, TLD reader stability and fading. The associated uncertainty is found to be about 25% for a 95% confidence interval (k = 2.025), which can be considered acceptable when taking into account the very low air kerma rates estimated. The LiF-700-H response to different energies and its relationship with climate changes over the calendar year are planned as future tasks. (Author)

  7. Leak detection device for control rod drive and detection method therefor

    International Nuclear Information System (INIS)

    Imasaki, Yoshio.

    1997-01-01

    The present invention provides a detection device for leak of cooling water from a sealed axial portion of control rod drives (CRD) disposed in a BWR type reactor and a monitoring method therefor. Namely, the CRD transfers rotation at the sealed axial portion and elevates/lowers a piston to insert/withdraw control rod into/from the reactor core. High pressure water is injected upon occurrence of scram to urge the piston upwardly thereby rapidly inserting the control rods. Leak detection pipelines are laid from the sealed axial portion. A flow glass is connected to the leak detection pipelines. Then, cooling water leaked from the sealed axial portion flows in the leak detection pipelines and flows into the flow glass. The flow rate of cooling water leaked from the sealed axial portion of the CRD can thus be detected by monitoring the flow glass. In addition, a flowmeter is connected to the leak detection pipelines, or the flowmeter and the flow glass are connected, and a flowmeter is connected downstream. Then, the flow rate of the leaked cooling water can be detected automatically. (I.S.)

  8. Validation of response simulation methodology of Albedo dosemeter

    International Nuclear Information System (INIS)

    Freitas, B.M.; Silva, A.X. da

    2016-01-01

    The Instituto de Radioprotecao e Dosimetria developed and runs a neutron TLD albedo individual monitoring service. To optimize the dose calculation algorithm and to infer new calibration factors, the response of this dosemeter was simulated. In order to validate this employed methodology, it was applied in the simulation of the problem of the QUADOS (Quality Assurance of Computational Tools for Dosimetry) intercomparison, aimed to evaluate dosimetric problems, one being to calculate the response of a generic albedo dosemeter. The obtained results were compared with those of other modeling and the reference one, with good agreements. (author)

  9. Unraveling thermoluminescence

    CERN Document Server

    Sunta, C M

    2014-01-01

    The book is essentially about the theoretical aspects of thermoluminescence (TL). It is structured and written in a manner which will specially be of interest to the fresh undergraduate students and young researchers planning to take up work in the field of thermoluminescence. It is arranged into seven chapters, starting with the historical growth of the subject of TL and an overview of the luminescence phenomena. A Physical model involving thermally disconnected deep traps is specially described. Using this model expressions are derived for the TL intensity growth curve, the supralinearity fa

  10. Calibration of a scintillation dosemeter for beta rays using an extrapolation ionization chamber

    International Nuclear Information System (INIS)

    Hakanen, A.T.; Sipilae, P.M.; Kosunen, A.

    2004-01-01

    A scintillation dosemeter is calibrated for 90 Sr/ 90 Y beta rays from an ophthalmic applicator, using an extrapolation ionization chamber as a reference instrument. The calibration factor for the scintillation dosemeter agrees with that given by the manufacturer of the dosemeter within ca. 2%. The estimated overall uncertainty of the present calibration is ca. 6% (2 sd). A calibrated beta-ray ophthalmic applicator can be used as a reference source for further calibrations performed in the laboratory or in the hospital

  11. Thermoluminescence of LaAlO3

    International Nuclear Information System (INIS)

    Morales H, A.; Zarate M, J.; Rivera M, T.; Azorin N, J.

    2015-10-01

    In this paper the thermoluminescent properties of doped lanthanum aluminate (LaAlO 3 ) with dysprosium ion (Dy) were studied. The thermoluminescence characteristics in the samples were obtained using an ultraviolet radiation of 220 nm. The LaAlO 3 :Dy samples were prepared by the modified Pechini method (Spray Dryer). The structural and morphological characterization was obtained by X-ray diffraction (XRD) and scanning electron microscopy (Sem) techniques respectively. The size particle composing the agglomerate was determined by Sem, agglomerated particles composed size of 2μm were observed. The thermoluminescence response of LaAlO 3 :Dy was compared with that obtained with the undoped sample. Thermoluminescence brightness curves of LaAlO 3 :Dy showed a peak centered at 185 grades C. Sensitivity of doped sample was greater, about 100 times compared with the undoped sample. Thermoluminescence response in function of the wavelength showed a maximum at 220 nm. Also the fading in thermoluminescence response was studied. (Author)

  12. Mexican gems as thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Azorin N, J.

    1979-01-01

    The possibility of using naturally ocurring mexican gems as thermoluminescent dosimeters (TLD) was investigated. Twelve types of gems were irradiated with X and gamma rays in order to determinate their dosimetric properties. Three of these gems showed favorable thermoluminescent characteristics compared with commercial thermoluminescent dosimeters. The plots of their thermoluminescent response as a function of gamma dose are straight lines on full log paper in the dose range 10 -2 to 10 2 Gy. The energy dependence is very strong to low energies of the radiation. Their fading was found to be about 5%/yr. and they may be annealed as reused without loss in sensitivity. Therefore, these gems can be used as X and gamma radiation dosimeters. (author)

  13. Measurements of environmental gamma-ray spectra using a multi-element TL dosemeter

    International Nuclear Information System (INIS)

    Furuta, Sadaaki; Boetter-Jensen, L.; Nielsen, S.P.

    1986-12-01

    A method to estimate the energy distribution and dose of environmental gamma radiation was developed using a multielement TL dosemeter. Experimentally obtained energy responses from a multi-element TL dosemeter with different kinds of filters were used to calculate the energy distribution and related dose by the SAND-II computer code. The code was originally developed to estimate the neutron flux using a multiple foil activation method. Measurements were made at several locations with the multi-element TL dosemeter and comparisons were made with results from a NaI(Tl) scintillation detector and a high-pressure ionization chamber. (author)

  14. Design, development and characterization studies of a large aperture high power Nd : glass rod amplifier stage

    International Nuclear Information System (INIS)

    Gopi, N.; Bapna, R.C.; Murali, C.G.; Narayan, B.S.; Dhareshwar, L.J.

    1992-01-01

    Laser-plasma interaction studies and experiments related to laser driven shocks as well as laser induced inertially confined thermonuclear fusion have resulted in an ever increasing demand for high brightness lasers capable of producing nanosecond pulse with energy of hundreds of kilo joules. High power Nd-glass laser chains with a master oscillator followed by a number of amplifier stages made up of rods, disks, slabs etc. are in operation in many leading laboratories in the world. This report describes the design, development and characterisation studies of a large aperture Nd:glass laser amplifier which is to be incorporated on line with the existing 40 J, 5 ns high power laser chain built for laser-plasma interaction and laser driven shock wave studies in the Laser and Plasma Technology Division. The development work described in this report discusses the design based on optimum material selection, optimisation of various sub components, ease of maintenance and smooth operation. The necessary operational electronics has also been described. The characterization studies mainly include measurement of spatial gain uniformity, thermally induced depolarization effects, and thermal relaxation studies. (author). 37 refs., 14 figs., 5 tabs

  15. EURADOS intercomparison 2012 for neutron dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Fantuzzi, E. [ENEA-Radiation Protection Institute, Bologna (Italy); Chevallier, M.A. [Institut de Radioprotection et de Surete Nucleaire (IRSN), Fontenay-aux-Roses (France); Cruz-Suarez, R. [International Atomic Energy Agency, Vienna (Austria); Luszik-Bhadra, M. [Physikalisch-Technische Bundesanstalt, Braunschweig (Germany); Mayer, S. [Paul Scherrer Institute (PSI), Villigen (Switzerland); Thomas, D.J. [National Physical Laboratory (NPL), Teddington (United Kingdom); Tanner, R. [Public Health England, Oxon (United Kingdom); Vanhavere, F. [SCK-CEN, Belgian Nuclear Research Centre, Mol (Belgium)

    2014-11-15

    EURADOS, within the work performed by Working Group 2 - Harmonization of Individual Monitoring in Europe, has started a self-sustained programme of regular intercomparisons and has successfully executed three intercomparisons for whole body photon dosemeters (IC2008, IC2010, IC2012) and one intercomparison for extremity dosemeters for photon and beta fields (IC2009). In 2012, the EURADOS intercomparison IC2012n was launched for personal neutron dosemeters routinely used to measure personal dose equivalent, Hp(10), in individual monitoring. No systems under development were allowed to participate. IC2012n was carried out by a EURADOS nominated Organization Group (OG) consisting of: Marie- Anne Chevallier (IRSN, F), Rodolfo Cruz-Suarez (IAEA, UN-Vienna), Marlies Luszik-Bhadra (PTB, D), Sabine Mayer (PSI, CH), David J. Thomas (NPL, UK), Rick Tanner (PHE, UK), Filip Vanhavere (SCK-CEN, B) led by a Coordinator, Elena Fantuzzi (ENEA, I). 31 participants registered for the comparison, with 34 dosimetry systems. In total 816 dosemeters were irradiated in selected neutron fields on an ISO slab phantom. The irradiations were performed at 2 European accredited laboratories which are both National Primary Metrology Laboratories for ionizing radiation: NPL (National Physical Laboratory, UK) and PTB (Physikalisch-Technische Bundesanstalt, D). All irradiations were carried out according to the irradiation plan developed by the OG. The overall results show that most, although not all, dosimetric systems perform acceptably well (within a factor of 2) for irradiations with a bare radionuclide source ({sup 252}Cf at 0 ), whilst more than half of the systems underestimate the reference value for irradiations from non-normal angles of incidence irradiations ({sup 252}Cf at 45 ) or for simulated workplace fields ({sup 252}Cf(D{sub 2}O) or {sup 252}Cf source behind a shadow cone). The performance for 250 keV mono-energetic neutron irradiations varies mainly reflecting the detection

  16. Thermoluminescence, a universal tool for dosimetry

    International Nuclear Information System (INIS)

    Barthe, J.

    1999-01-01

    Radio-thermoluminescence is a transient light emission occurring when certain substances, which have been exposed to ionizing radiation, are heated. Under some conditions the quantity of emitted light is proportional to the absorbed dose. This linearity allows thermoluminescence to be used in various fields such as dosimetry, geology and age determination. Dosimeters based on thermoluminescence are passive, they do not require a source of energy and they can sustain high temperatures and high levels of irradiation. Thermoluminescence has been applied for determining the dose absorbed by meteorites and lunar rocks

  17. On the calibration of radiotherapy dosemeters in Australia

    International Nuclear Information System (INIS)

    Huntley, R.; Kotler, L.; Webb, D.

    2000-01-01

    Full text: Dosemeters for external beam radiotherapy are calibrated in Australia by ARPANSA, against the national primary standards of exposure and absorbed dose. The primary standards are free air chambers for exposure at low and medium energy X-rays, a graphite cavity chamber for exposure at 60 Co, and a graphite calorimeter for absorbed dose at 60 Co and high energy (MV) X -rays. Radiotherapy dosemeters are calibrated against these standards using a well documented formalism to provide calibration factors suitable for use with dosimetry protocols. A dosemeter usually comprises an ionization chamber connected to an independent electrometer. These are calibrated separately if possible. A combined calibration factor is reported together with the electrometer calibration factor (sensitivity). The dosimetry protocol used in radiotherapy centres in Australia and New Zealand is currently the simplified version of the IAEA TRS277 protocol, published by the New Zealand NRL and recommended by the ACPSEM. This protocol requires the use of an exposure or air kerma calibration factor at 60 Co (Nx or Nk) to evaluate the absorbed dose to air calibration factor N D . The chamber is then placed in a water phantom with its centre displaced from the reference point by p eff . ARPANSA can also supply calibration factors in absorbed dose to water (N D,w ), as required as input to the new IAEA CoP. If an absorbed dose to water calibration factor is used by the radiotherapy centre, the chamber should be placed with its centre at the reference point in the water phantom. ARPANSA has for some years coordinated the participation of Australian radiotherapy centres in the IAEA TLD Quality Audit service. Note that this service does not represent a calibration and should not be referred to as such. The only calibration is that provided by ARPANSA for a reference dosemeter at each radiotherapy centre. As soon as the ANSTO SSDL is operational, calibrations of reference dosemeters will also be

  18. On the chemical variability of Middelburg glass beads and rods

    International Nuclear Information System (INIS)

    Karklins, K.; Kottman, J.; Hancock, R.G.V.; Sempowski, M.L.; Nohe, A.W.; Moreau, J.-F.; Aufreiter, S.; Kenyon, I.

    2001-01-01

    Forty-three glass samples from a late 16th-early 17th century, glass beadmaking house in Middelburg, the Netherlands, were selected for maximum colouring variability, including plain and multi-coloured varieties. The glass chemistries were quite diverse, within each colour grouping. For each single colour of glass, anticipated colouring elements (copper for turquoise blue, cobalt for dark blue, manganese for rose, and tin for white) were used, with the exception of two beads that were opacified wih antimony rather than with tin. Multi-coloured glass glasses (chevron beads) produced chemistries that match the mixing of the different coloured glasses. In some cases, low relative amounts of some inter-mixed glasses were not detectable against the composition of the major glass component. (author). 16 refs., 3 tabs

  19. C-program LINOP for the evaluation of film dosemeters by linear optimization. User manual

    International Nuclear Information System (INIS)

    Kragh, P.

    1995-11-01

    Linear programming results in an optimal measuring value for film dosemeters. The Linop program was developed to be used for linear programming. The program permits the evaluation and control of film dosemeters and of all other multi-component dosemeters. This user manual for the Linop program contains the source program, a description of the program and installation and use instructions. The data sets with programs and examples are available upon request. (orig.) [de

  20. Thermoluminescence dating of pottery

    International Nuclear Information System (INIS)

    Higashimura, Takenobu; Ichikawa, Yoneta.

    1978-01-01

    This report is divided into two parts. The first half describes on the history of thermoluminescence dating, and the latter half, the principle and measurement examples. It was in late 1955 that the measurement of radiation dose using thermoluminescence began. The method to thermoluminescence dating was developed when it was found that most natural stones emit the thermoluminescence. About Greek earthen wares, the study of which was presented in 1961 by G. Kennedy of University of California, the dating was able to be made within the standard deviation of 10%. Since then, this dating method progressed rapidly, and a number of laboratories are now forwarding the investigation. In the samples of natural materials, intensity of thermoluminescence I is proportional to natural radiation dose D which has been absorbed by the samples, i.e. I = kD, where k is the susceptibility of thermoluminescence of the samples. Since k is different in each sample, D can be determined by irradiating the sample with β or γ ray of known dose D 0 , measuring its luminescence I 0 , and eliminating k through these two equations, because i 0 = kD 0 . Next, if t is assumed to be the time passed since a pottery was made, D is expressed as Rt, where R is the natural radiation dose per year absorbed by the pottery. Thus t is determined if R is known. The report describes on the method of measuring R. As an example, the results of measurement of the potteries excavated at Iwakura remains, Yorikura, Taishaku-kyo, are listed. Results by 14 C dating are also described for reference. (Wakatsuki, Y.)

  1. Method for detecting radiation dose utilizing thermoluminescent material

    International Nuclear Information System (INIS)

    Miller, S.D.; McDonald, J.C.; Eichner, F.N.; Durham, J.S.

    1992-01-01

    The amount of ionizing radiation to which a thermoluminescent material has been exposed is determined by first cooling the thermoluminescent material and then optically stimulating the thermoluminescent material by exposure to light. Visible light emitted by the thermoluminescent material as it is allowed to warm up to room temperature is detected and counted. The thermoluminescent material may be annealed by exposure to ultraviolet light. 5 figs

  2. Development of radiophotometric dosemeters of high sensitivity using plastic scintillators as light intensifiers

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Hamada, M.M.

    1987-01-01

    The use of rectangular plates of plastic scintillators as film holders in conventional photographic dosemeters is reported. The efficiency of their use as light converters for increase the sensitivity of these dosemeters are studied. (M.A.C.) [pt

  3. Towards Luminescence Dating Of Mosaic Glass

    Science.gov (United States)

    Galli, A.; Martini, M.; Sibila, E.; Villa, I.

    The possibility of dating archaeological glass by means of luminescent techniques has been investigated in recent years, despite the difficulties of this application, mainly linked to the amorphous structure of the material. We focused in particular on mosaic glass, after the encouraging results obtained on byzantine and medieval samples. Further studies were devoted to the comprehension of the luminescent mechanisms in silica glasses, and to the investigation of the relationships between luminescence, colouring or opacifier ions and crystalline phase of the vitreous matrix. The results of a study on the dosimetric characteristics of thermoluminescence (TL) and Optically Stimulated Luminescence (OSL) of a few medieval blue-green mosaic glasses from the San Lorenzo church (Milan) are presented, and the experimental protocols established to identify their suitability for dating are discussed.

  4. Personnel photon dosimeter on the base of TLD sup n sup a sup t LiF

    CERN Document Server

    Kaskanov, G Y

    2003-01-01

    A personnel photon dosemeter on the basis of thermoluminescence detectors (TLD) sup n sup a sup t LiF is described. Experimental responses of the dosemeters in the unit of individual equivalent dose H sub p (10) for energy of photons 59.5, 120, 662, and 1250 keV are presented. It is shown that the dosemeter allows one to measure H sub p (10) with admissible uncertainty in the energy range from 60 to 1250 keV.

  5. Thermoluminescence properties of the Cu-doped lithium potassium borate glass

    International Nuclear Information System (INIS)

    Aboud, Haydar; Wagiran, H.; Hussin, R.; Ali, Hassan; Alajerami, Yasser; Saeed, M.A.

    2014-01-01

    Characteristics of lithium potassium borate glasses with various copper concentrations are reported. The glasses were prepared by the melt quenching method and irradiated with photons to doses in the 0.5–4.0 Gy range. Glowing curves, dose response curves, reproducibility of the response, dose threshold, thermal fading and optical bleaching were studied

  6. Neutron dosemeter responses in workplace fields and the implications of using realistic neutron calibration fields

    International Nuclear Information System (INIS)

    Thomas, D.J.; Horwood, N.; Taylor, G.C.

    1999-01-01

    The use of realistic neutron calibration fields to overcome some of the problems associated with the response functions of presently available dosemeters, both area survey instruments and personal dosemeters, has been investigated. Realistic calibration fields have spectra which, compared to conventional radionuclide source based calibration fields, more closely match those of the workplace fields in which dosemeters are used. Monte Carlo simulations were performed to identify laboratory systems which would produce appropriate workplace-like calibration fields. A detailed analysis was then undertaken of the predicted under- and over-responses of dosemeters in a wide selection of measured workplace field spectra assuming calibration in a selection of calibration fields. These included both conventional radionuclide source calibration fields, and also several proposed realistic calibration fields. The present state of the art for dosemeter performance, and the possibilities of improving accuracy by using realistic calibration fields are both presented. (author)

  7. A personal dosemeter of radon daughter products

    International Nuclear Information System (INIS)

    Burian, I.; Richter, M.; Ochec, J.; Dohnal, J.

    1990-01-01

    The OD 88 personal dosemeter for radon daughter products uses an alpha particle track detector and is of active type, with forced air aspiration through the detecting system. The pumping device consists of a miniature membrane pump and an electronic speed controller. The air flow rate is little affected by normal changes in supply voltage and/or pneumatic resistance of the pump circuit. The dosemeter is embedded in a miner's lamp and is fed from its accumulator battery. The pumping device is activated as the lamp is taken out of the charger and deactivated as it is slid into the charger again. The dosemeter head accommodates an AFPC filter on which the radon daughter products deposit. The energy of the alpha particles is reduced by an air layer and a Makrofol decelerating foil. The particles hitting the detector give rise to defects with an intensity proportional to the total linear stopping power at the moment of incidence; ideal tracks 10 μm in size require incident energy of 1.1 to 1.3 MeV. The detector is replaced in a month and worked up in an etching vessel, making use of the higher solubility of the detecting layer of the foil in NaOH at the spot of the alpha particle incidence. (P.A.). 4 figs., 17 refs

  8. Studies on γ-ray induced structural changes in Nd{sup 3+} doped lead alumino silicate glasses by means of thermoluminescence for dosimetric applications in high dose ranges

    Energy Technology Data Exchange (ETDEWEB)

    Sundara Rao, M. [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, Guntur, A.P. (India); Gandhi, Y. [Department of Physics, Kakani Venkata Ratnam College, Nandigama 521 185, A.P. (India); Sanyal, Bhaskar [Food Technology Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400 085 (India); Bhargavi, K. [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, Guntur, A.P. (India); Piasecki, M. [Institute of Physics, J. Dlugosz University, Ul. Armii Krajowej 13/15, 42-201 Czestochowa (Poland); Veeraiah, N., E-mail: nvr8@rediffmail.com [Department of Physics, Acharya Nagarjuna University, Nagarjuna Nagar 522 510, Guntur, A.P. (India)

    2014-12-15

    Graphical abstract: TL glow curves of PbO–SiO{sub 2}:Nd{sup 3+} glasses mixed with different concentrations of Al{sub 2}O{sub 3} exposed to γ-rays of dose 5.0 kGy. - Highlights: • TL studies of Nd{sup 3+} ions doped lead alumino silicate glasses were carried out. • Highest TL output was observed in the glasses mixed with 10 mol% of Al{sub 2}O{sub 3}. • Different mechanisms responsible for TL emission were discussed. • Near linearity of the dose response was observed in the dose range of 1.0–3.0 kGy. • These glasses may be useful as dosimeters in processing perishable food commodities. - Abstract: Thermoluminescence (TL) studies on PbO–Al{sub 2}O{sub 3}–SiO{sub 2}:Nd{sup 3+} glasses mixed with varying concentrations of Al{sub 2}O{sub 3} exposed to γ-rays of dose in the range 0–5.0 kGy were carried out. The TL emission exhibited a dosimetric peak at about 185 °C. The TL output under the glow peak increased with increasing Al{sub 2}O{sub 3} and also with the γ-ray dose. The mechanisms responsible for TL emission and enhancement of TL output with increase in the concentration of Al{sub 2}O{sub 3} were quantitatively discussed in terms of induced structural defects in the vicinity of Nd{sup 3+} ions due to interaction of γ-rays with the glass network in the scenario of varying concentration of Al{sub 2}O{sub 3}. The dose response of these glass samples exhibited linear behavior in the medium dose range viz., 1.0–3.0 kGy. Finally, it is concluded that the glass containing the highest concentration of Al{sub 2}O{sub 3} exhibits high TL output and such glasses are useful for dosimetry in the range 1.0–3.0 kGy and hence these glasses may be useful for dosimetry in such high range of doses required for commercial radiation processing of perishable food commodities to extend their shelf-lives.

  9. Assessment of eye lens doses in interventional radiology: a simulation in laboratory conditions

    International Nuclear Information System (INIS)

    Cemusova, Z.; Ekendahl, D.; Judas, L.

    2016-01-01

    As workers in interventional radiology belong to one of the most occupationally exposed groups, methods for sufficiently accurate quantification of their external exposure are sought. The objective of the authors' experiment was to investigate the relations between eye lens dose and H p (10), H p (3) or H p (0.07) values measured with a conventional whole-body personal thermoluminescence dosemeter (TLD). Conditions of occupational exposure during common interventional procedures were simulated in laboratory. An anthropomorphic phantom represented a physician. The TLDs were fixed to the phantom in different locations that are common for purposes of personal dosimetry. In order to monitor the dose at the eye lens level during the exposures, a special thermoluminescence eye dosemeter was fixed to the phantom's temple. Correlations between doses measured with the whole-body and the eye dosemeters were found. There are indications that personnel in interventional radiology do not need to be unconditionally equipped with additional eye dosemeters, especially if an appropriate whole-body dosimetry system has been already put into practice. (authors)

  10. Skin bridge versus rod colostomy in children - comparison between complications.

    Science.gov (United States)

    Askarpour, Shahnam; Peyvasteh, Mehran; Changai, Bahram; Javaherizadeh, Hazhir

    2012-10-01

    Due to economic problems, sigmoid loop colostomy using glass rod may cause problems for our patients for finding glass rod and several visits. The aim of the study was to compare rod versus skin bridge colostomy. In this study, 42 cases who are candidate for colostomy were included. Cases were randomly placed in skin bridge and rod colostomy group. Independent sample t-test and Chi-square were used for comparison. SPSS version 16.0 (SPSS Inc, Chicago, IL, USA) was used for analysis. Of 42 cases, 20 were male and 22 were female. Hirschsprung's disease was the indication of colostomy in 33 cases. In nine cases, imperforate anus was the indication of colostomy. Mean time of surgery was 79.4 and 82.5 minute for the rod and skin bridge group respectively (P>0.05). Retraction was seen in 2 case of rod group, and no case of skin bridge group. Prolapse was seen in 2 (9.5%) case of rod group and 1(4.7%) case in skin bridge. There were no reports of necrosis, stenosis, and hernia in both groups. In the skin bridge group the rates of complications were lower but the groups are too small for statistical analysis. Colostomy with a skin bridge method may decrease number of revision and expenses and may be appropriate option. Sigmoid loop colostomy using skin bridge flap may be appropriate choice in developing country. Another study with more samples is recommended to better comparison of Skin Bridge versus rod colostomy.

  11. Dosimetric Characteristics of a LKB:Cu,Mg Solid Thermoluminescence Detector

    International Nuclear Information System (INIS)

    Alajerami Yasser Saleh Mustafa; Hashim Suhairul; Ramli Ahmad Termizi; Saleh Muneer Aziz; Kadir Ahmad Bazlie Bin Abdul; Saripan, Mohd. Iqbal

    2013-01-01

    We present the main thermoluminescence characteristics of a newly borate glass dosimeter modified with lithium and potassium carbonate (LKB) and co-doped with CuO and MgO. An enhancement of about three times has been shown with the increment of 0.1mol% MgO as a co-dopant impurity. The effects of dose linearity, storage capacity, effective atomic number and energy dose response are studied. The proposed dosimeter shows a simple glow curve, good linearity up to 10 3 Gy, close effective atomic number and photon energy independence. The current results suggest using the proposed dosimeter in different dosimetric applications

  12. Micro-dosemeter instrument (MIDN) for assessing risk in space

    International Nuclear Information System (INIS)

    Pisacane, V. L.; Dolecek, Q. E.; Malak, H.; Cucinotta, F. A.; Zaider, M.; Rosenfeld, A. B.; Rusek, A.; Sivertz, M.; Dicello, J. F.

    2011-01-01

    Radiation in space generally produces higher dose rates than that on the Earth's surface, and contributions from primary galactic and solar events increase with altitude within the magnetosphere. Presently, no personnel monitor is available to astronauts for real-time monitoring of dose, radiation quality and regulatory risk. This group is developing a prototypic instrument for use in an unknown, time-varying radiation field. This micro-dosemeter-dosemeter nucleon instrument is for use in a space-suit, spacecraft, remote rover and other applications. It provides absorbed dose, dose rate and dose equivalent in real time so that action can be taken to reduce exposure. Such a system has applications in health physics, anti-terrorism and radiation-hardening of electronics as well. The space system is described and results of ground-based studies are presented and compared with predictions of transport codes. An early prototype in 2007 was successfully launched, the only solid-state micro-dosemeter to have flown in space. Published by Oxford Univ. Press on behalf of the US Government 2011. (authors)

  13. Investigation of thermoluminescence properties of mobile phone screen displays as dosimeters for accidental dosimetry

    Science.gov (United States)

    Mrozik, Anna; Marczewska, B.; Bilski, P.; Kłosowski, M.

    2014-11-01

    The rapid assessment of the radiation dose after unexpected exposure is a task of accidental dosimetry. In case of a radiological accident glasses originating from mobile phone screens, placed usually near the human body, could be used as emergency thermoluminescent (TL) personal dosimeters. The time between irradiation and TL readout is crucial and therefore preparation of the mobile phone screens and their readout conditions should be optimized. The influence of the samples etching, bleaching and selection of the optical filters based on measurement of the emission spectrum of irradiated glass samples during heating for different types of mobile phones were the subjects of our investigation. Obtained results showed that glasses extracted from different brands of mobile phones have different dosimetric properties but all of them give a luminescent signal which can be used to calculate the dose.

  14. Operational trials of single- and multi-element CR-39 dosemeters for the DIDO and PLUTO reactors at the Harwell Laboratory

    International Nuclear Information System (INIS)

    Gallacher, G.G.; Perks, C.A.

    1993-01-01

    Single- and multi-element CR-39 dosemeters, developed at the Harwell Laboratory, and a commercially available multi-element CR-39 dosemeter (obtained from Track Analysis Systems Ltd), were evaluated for their potential as neutron dosemeters for personnel working at Harwell Laboratory's research reactors. Owing to the angular dependence of the CR-39 (processed using electrochemical etching), the single-element dosemeter was found to be impractical. Consequently, a multi-element dosemeter was developed, which consisted of a cube of side 36 mm with CR-39 elements (also processed using electrochemical etching) attached to each of the sides. Although this dosemeter was technically suitable for this type of dosimetry, it was considered to be unacceptably bulky in personnel trials. The commercially available CR-39 dosemeter tested was much smaller (the CR-39 was only chemically etched) and this was considered to be acceptable as a personnel dosemeter. In addition, trials with personnel working at active handling glove boxes indicated that single-element dosemeters might be adequate, but further work would be needed to verify this. (author)

  15. Characteristic of the dosemeter used by the service of film dosimetry in the CPHR

    International Nuclear Information System (INIS)

    Capote Ferrera, E.A.; Diaz Bernal, E.; Manzano de Armas, J.F.

    1998-01-01

    Characteristic of the dosemeter used by the service of dosimetria filmica of the CPHR.The use of film dosemeter brings I get the aquisition first it gives data that it includes its evaluation regarding the calibration magnitude and second the interpretation of these data in terms of the magnitude of personal dose, such as the equivalent dose (ICRU 47) or to the effective dose (ICRP 60). presently work the results of the study of the characteristics of the dosemeters are picked up

  16. Production of thermoluminescent dosemeters based on MgB4O7: Dy and MgB4O7: Tm

    International Nuclear Information System (INIS)

    Souza, Luiza Freire de; Souza, Divanizia N.

    2013-01-01

    The thermoluminescent dosimetry (TL) is a well-established technique for the detection of ionizing radiation in hospitals, clinics, and industrial establishments where there is the need to quantify the radiation. For this practice is require the use phosphors which are sensitive to radiation. Some phosphors are already commonly used in this practice, for example, TLD-100 (LiF: Mg, Ti), CaSO 4 :Tm and CaSO 4 :Dy. A compound that was most recently introduced in dosimetry and has many advantageous features to detect neutrons, electrons and gamma is the magnesium tetraborate (MgB 4 O 7 ), but the undoped material is not good for dosimetry, since signal does not show satisfactory thermoluminescence. The present work presents the analysis of the compound MgB 4 O 7 when doped with rare earth elements, thulium (Tm) and dysprosium (Dy). The production of MgB 4 O 7 : Dy and MgB 4 O 7 : Tm occurred under acidic conditions. Following the process of crystal growth, several tests were made on phosphors produced to verify the quality of materials as TL dosimeter. Initially, was made the identification of the crystalline phases found in the material, using the technique of X-ray diffractometry, and then were evaluated and compared the TL emission curves of the crystals with two different types of dopants, to this, the samples were irradiated with different radiation sources: 137 Cs (0,66 MeV), 60 Co (1.25 MeV) and X-rays (0.41 MeV) and based on the results was evaluated the energy dependence of phosphors. Another characteristic analyzed, was the decay of TL signal for the material (fading). The results show that the material can be an excellent TL dosimeter when doped with rare earth elements Dy and Tm. (author)

  17. Energy dependence of pMOS dosemeters

    International Nuclear Information System (INIS)

    Savic, Z.; Stankovic, S.; Kovacevic, M.; Petrovic, M.

    1996-01-01

    The results are presented of experimental work and numerical simulations of the energy response for pMOS dosimetric transistors in their custom packages. Specially produced radiation soft pMOS transistors were used in this experimental work. The irradiation of pMOS dosemeters was done using 60 Co and 137 Cs sources, a dosimetric X ray unit, and a radiotherapeutic linear accelerator in the range of photon energies from 21 keV to 8 MeV. The results show that package geometry and materials can significantly affect and smooth the energy dependence of pMOS transistors and that in custom transistor packages they are not tissue-equivalent dosemeters. Their response in the photon energy range of 45 to 250 keV is significantly larger than it should be (maximum dose enhancement factor can be as high as 8) and some energy compensation techniques must be used in order to fulfill the requirements of corresponding standards. (Author)

  18. Energy dependence of pMOS dosemeters

    Energy Technology Data Exchange (ETDEWEB)

    Savic, Z. [Military Technical Institute, Belgrade (Yugoslavia); Stankovic, S.; Kovacevic, M.; Petrovic, M. [Institute of Nuclear Sciences, Belgrade (Yugoslavia). Radiation Protection Dept.

    1996-10-01

    The results are presented of experimental work and numerical simulations of the energy response for pMOS dosimetric transistors in their custom packages. Specially produced radiation soft pMOS transistors were used in this experimental work. The irradiation of pMOS dosemeters was done using {sup 60}Co and {sup 137}Cs sources, a dosimetric X ray unit, and a radiotherapeutic linear accelerator in the range of photon energies from 21 keV to 8 MeV. The results show that package geometry and materials can significantly affect and smooth the energy dependence of pMOS transistors and that in custom transistor packages they are not tissue-equivalent dosemeters. Their response in the photon energy range of 45 to 250 keV is significantly larger than it should be (maximum dose enhancement factor can be as high as 8) and some energy compensation techniques must be used in order to fulfill the requirements of corresponding standards. (Author).

  19. In vivo thermoluminescent dosimetry in studies of helicoid computed tomography and excretory urogram; Dosimetria termoluminiscente In vivo en estudios de tomografia computada helicoidal y urograma excretor

    Energy Technology Data Exchange (ETDEWEB)

    Cruz C, D.; Azorin N, J. [UAM-I, 09340 Mexico D.F. (Mexico); Saucedo A, V.M.; Barajas O, J.L. [Unidad de Especialidades Medicas, Secretaria de la Defensa Nacional, 11500 Mexico D.F. (Mexico)

    2005-07-01

    The dosimetry is the field of measurement of the ionizing radiations. It final objective is to determine the 'absorbed dose' for people. The dosimetry is vital in the radiotherapy, the radiological protection and the treatment technologies by irradiation. Presently work, we develop 'In vivo' dosimetry, in exposed patients to studies of helical computed tomography and excretory urogram. The dosimetry 'in vivo' was carried out in 20 patients selected aleatorily, for each medical study. The absorbed dose was measured in points of interest located in crystalline, thyroid, chest and abdomen of each patient, by means of thermoluminescent dosemeters (TLD) LiF: Mg,Cu,P + Ptfe of national fabrication. Also it was quantified the dose in the working area. (Author)

  20. Thermoluminescent system for low temperatures

    International Nuclear Information System (INIS)

    Rosa, L.A.R. da; Caldas, L.V.E.; Leite, N.G.

    1988-09-01

    A system for measurements of the thermoluminescent glow curve, the thermoluminescent emission spectrum and the optical absorption spectrum of solid samples, from liquid nitrogen temperature up to 473 K, is reported. A specially designed temperature programmer provides a linear heating of the sample at a wide range of selectable heating rates, as also long term steady-state temperatures for annealing and isothermal decay studies. The system operates at a pressure of 1.33 x 10 -3 Pa. Presently it is being used for lithium fluoride low temperature thermoluminescent studies. (author) [pt

  1. Development of a transfer instrument for neutron dosimetry intercomparison

    International Nuclear Information System (INIS)

    Greene, D.; Miles, J.

    1974-01-01

    Comparisons are reported for fast neutron dosemeters which were designed to be transportable so as to enable intercomparisons between institutions using neutrons for radiotherapy or radiobiology. The systems considered are : 1) the ferrous sulphate dosemeter, 2) the lithium fluoride thermoluminescent dosemeter, 3) ionization chambers with various walls and gases. Work on photographic film dosimetry indicated that the system was not suitable and was not pursued. The sources used were 60 Co, the cyclotron at Hammersmith Hospital in London and 252 Cf

  2. Accuracy of megavolt radiation dosimetry using thermoluminescent lithium fluoride.

    Science.gov (United States)

    Rudén, B I; Bengtsson, L G

    1977-04-01

    The relative light output per Gy in polystyrene for roentgen beams of 6 and 42 MV and electrons between 2.2 and 34.5 MeV relative to 60Co gamma radiation is reported for different kinds of LiF dosemeters. The distribution of the absorbed dose inside a 0.25 and 0.4 mm thick LiF-teflon disc surrounded by polystyrene and irradiated with 60Co, 42 MV roentgen radiation and 39 MeV electrons was measured using 0.01 and 0.02 mm thick Lif-teflon discs. The measurements show that the absorbed dose distribution in the dosemeter depends on the energy of the radiation. When flat dosemeters were used, differences between the signals measured at the two orientations possible during read-out could easily amount to several per cent, and for this reason 0.4 mm and 0.5 mm LiF-Teflon discs were not trusted when the highest accuracy was required. The cavity theory by Burlin does not account for the phenomena caused by differences in electron scattering properties of the dosemeter and the phantom material. Some suggestions are presented for a different cavity theory for flat dosemeters dealing also with these phenomena. It describes the results to about the same degree of approximation as the Burlin theory, and fails to explain the observed energy dependence for electrons.

  3. Study and determination of thermoluminescent properties of a new dosemeter based on α -Al2O3:C

    International Nuclear Information System (INIS)

    Rivera Montalvo, T.

    1993-01-01

    Study and determination of the thermoluminescent properties of a new dosimeter based on α -Al 2 O 3 :C. Dosimetric characteristics studied were: glow curve structure, sensitivity, reproducibility, TL response as function of dose, and fading. Each experimental data point represents the average value of at least five measurements. The new detectors exhibited two peaks in its glow curve at 383 and 493 K. The low temperature peak decayed at room temperature in a few hours. Sensitivity of α -Al 2 O 3 :C detectors to gamma radiation per unit of absorbed dose was 3 times that of LiF:Mg, Cu, P and 11 times that of CaSO 4 : Dy + PTFE; the last two detectors possess characteristics which make them appropriate for diverse applications in the field of radiation dosimetry, specially for environmental measurements. (Author)

  4. Verification of Ca F2:Mn type of dosemeters for personal dosimetry purposes

    International Nuclear Information System (INIS)

    Misovic, M.; Boskovic, Z.; Spasic-Jokic, V.

    1995-01-01

    Verification results of CaF2:Mn type of dosemeters for personal dosimetry purposes are presented in this paper. Tree types of irradiations are proceeded due to verification of relevant features of TLD. It is concluded that mentioned type of dosemeter can be used for purpose of personal dosimetry. (author)

  5. Dosimetric properties of dysprosium doped calcium magnesium borate glass subjected to Co-60 gamma ray

    Energy Technology Data Exchange (ETDEWEB)

    Omar, R. S., E-mail: ratnasuffhiyanni@gmail.com; Wagiran, H., E-mail: husin@utm.my; Saeed, M. A. [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 Johor Bahru (Malaysia)

    2016-01-22

    Thermoluminescence (TL) dosimetric properties of dysprosium doped calcium magnesium borate (CMB:Dy) glass are presented. This study is deemed to understand the application of calcium as the modifier in magnesium borate glass with the presence of dysprosium as the activator to be performed as TL dosimeter (TLD). The study provides fundamental knowledge of a glass system that may lead to perform new TL glass dosimetry application in future research. Calcium magnesium borate glass systems of (70-y) B{sub 2}O{sub 3} − 20 CaO – 10 MgO-(y) Dy{sub 2}O{sub 3} with 0.05  mol % ≤ y ≤ 0.7  mol % of dyprosium were prepared by melt-quenching technique. The amorphous structure and TL properties of the prepared samples were determined using powder X-ray diffraction (XRD) and TL reader; model Harshaw 4500 respectively. The samples were irradiated to Co-60 gamma source at a dose of 50 Gy. Dosimetric properties such as annealing procedure, time temperature profile (TTP) setting, optimization of Dy{sub 2}O{sub 3} concentration of 0.5 mol % were determined for thermoluminescence dosimeter (TLD) reader used.

  6. Fundamental study on the characteristics of a radiophotoluminescence glass dosemeter with no energy compensation filter for measuring patient entrance doses in cardiac interventional procedures

    International Nuclear Information System (INIS)

    Kato, Mamoru; Chida, Koichi; Moritake, Takashi; Koguchi, Yasuhiro; Sato, Tadaya; Kadowaki, Ken; Oosaka, Hajime; Tosa, Tetsuo

    2014-01-01

    Cardiac interventional procedures have been increasing year by year. However, radiation skin injuries have been still reported. There is a necessity to measure the patient entrance skin dose (ESD), but an accurate dose measurement method has not been established. To measure the ESD, a lot of radiophotoluminescence dosemeters (RPLDs) provide an accurate measurement of the direct actual ESD at the points they are arrayed. The purpose of this study was to examine the characteristics of RPLD to measure the ESD. As a result, X-ray permeable RPLD (with no tin filter) did not interfere with the percutaneous coronary intervention procedure. The RPLD also had good fundamental performance characteristics. Although the RPLD had a little energy dependence, it showed excellent dose and dose-rate linearity, and good angular dependence. In conclusion, by calibrating the energy dependence, RPLDs are useful dosemeter to measure the ESD in cardiac intervention. (authors)

  7. Preparation of thermoluminescent materials

    International Nuclear Information System (INIS)

    1976-01-01

    Thermoluminescent materials have been found to be suitable for measuring long term exposures to low level ionizing radiation. Oxyhalides of lanthanum, gadolinium and yttrium, including the oxychlorides and oxybromides are activated with terbium and have been found to be most efficient oxygendominated phosphors having thermoradiant efficiencies with excitation by low level ionizing radiation. Thermoluminescence response increases when the previous materials have hafnium and zirconium additives

  8. Development of new dosemeters of CaSO_4

    International Nuclear Information System (INIS)

    Santos, Ana Gabryele M. dos; Junot, Danilo O.; Souza, Divanizia N.

    2016-01-01

    Thermoluminescent dosimeters (TLDs) are widely used in in vivo dosimetry because of their simple production, viability for dose determination in difficult geometries and their reusability. The motivation of this work was to produce CaSO_4 crystals doped with thulium (Tm) and analyze their thermoluminescent (TL) properties. The crystals were produced by a variation of slow evaporation route, and TL curves showed emission peaks centered around 190 ° C and 320 ° C after irradiation with a "9"0Sr /"9"0Y source. Thermoluminescent properties such as linearity, reproducibility and fading were evaluated. The CaSO_4:Tm samples produced proved feasible for dosimetry purposes. (author)

  9. Eurados trial performance test for personal dosemeters for external beta radiation

    DEFF Research Database (Denmark)

    Christensen, P.; Bordy, J.M.; Ambrosi, P.

    2001-01-01

    On the initiative of the European Dosimetry Group (EURADOS) action group 'Harmonisation and Dosimetric Quality Assurance in Individual Monitoring for External Radiation' a trial performance test for whole-body and extremity personal dosemeters broadly representative of those in use in the EU...... the results obtained from the exercise. In particular, based on the replies to a questionnaire issued to each participant, the results are analysed in relation to important design characteristics of the dosemeters taking part in the test....

  10. Evaluation of the glow curves of a new glass matrix; Avaliação das curvas de emissão termoluminescente de uma nova matriz vítrea

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Nathália S.; Souza, Samara P.; Ferreira, Pâmela Z.; Dantas, Noelio O.; Silva, Anielle C.A.; Neves, Lucio P.; Perini, Ana P., E-mail: anapaula.perini@ufu.br [Universidade Federal de Uberlândia (INFIS/UFU), Uberlândia, MG (Brazil). Instituto de Física; Caldas, Linda V.E. [Instituto de Pesquisas Energéticas e Nucleares (IPEN/CNEN-SP), São Paulo, SP (Brazil); Carrera, Betzabel N.S.; Watanabe, Shigueo [Universidade de São Paulo (IF/USP), São Paulo, SP (Brazil). Instituto de Física

    2017-07-01

    Thermoluminescence is a dosimetric technique with may be used to personal, clinical, environmental and high doses. In this work a new glass matrix, with nominal composition of 20Li{sub 2}CO{sub 3}.10Al{sub 2}O{sub 3}.25BaO.45B{sub 2}O{sub 3} (mol%), was studied by the thermoluminescence technique. The glow curves was be analyzed, after the irradiation of this glass matrix with high doses. The results showed that this new glass matrix has a temperature peak in 260°C, which is ideal for dosimetry applications. (author)

  11. TL response of LiF: Mg, Cu, P dosemeters in function of the photon energy

    International Nuclear Information System (INIS)

    Gonzalez, P.R.; Azorin, J.; Furetta, C.

    2005-01-01

    In this work the obtained results of studying the response of dosemeters of LiF: Mg, Cu, P + Ptfe, developed in the ININ, commercial dosemeters GR200A (LiF: Mg, Cu, P) of chinese manufacture and TLD-100 (LiF: Mg, Ti) US are presented, when irradiating them with X rays 16, 24, 34.5, 42, 100 and 145 keV of effective energy, with gamma rays of 662 keV ( 137 Cs) and 1252 keV ( 60 Co). The results are presented in function of the sensitivity that the dosemeters showed normalized to gamma radiation of 60 Co. It was appreciated that the dosemeters more equivalent to the tissue, they were those of LiF: Mg, Cu, P + Ptfe, while the less equivalent ones were the TLD-100. (Author)

  12. Studies of aluminium nitride ceramics for application in UV dosimetry

    DEFF Research Database (Denmark)

    Trinkler, L.; Bøtter-Jensen, L.; Christensen, P.

    2000-01-01

    The study is reported of the ceramic material AlN-Y2O3 as a potential luminescence dosemeter for the detection of UV radiation. Both the thermoluminescence and the optically stimulated luminescence properties of the material have been studied after exposure to UV radiation and compared with those...... of the widely used dosemeter material Al2O3:C. It has been shown that AlN-Y2O3 ceramics exhibit three orders of magnitude higher sensitivity to UV radiation than does Al2O3,:C over a broad spectral region. The thermoluminescence from AlN-Y2O3 is characterised by linear dose dependence over a wide range....... The fading characteristics of the UV-induced thermoluminescence and optically stimulated luminescence signals with storage time at room temperature were found to be a drawback, but still lower than those induced after exposure to ionising radiation....

  13. Thermoluminescence: its understanding and applications

    International Nuclear Information System (INIS)

    Nambi, K.S.V.

    1977-01-01

    The phenomenon of thermoluminescence is treated exhaustively both mathematically and physically. All the possible applications of the phenomenon are briefly described and an optimistic picture of the future trends in this field is projected. A review of latest information on thermoluminescence instrumentation and phosphors is also included [pt

  14. New thermoluminescence techniques for mineral exploration

    International Nuclear Information System (INIS)

    Levy, P.W.; Holmes, R.J.; Ypma, P.J.; Chen, C.C.; Swiderski, H.S.

    1977-01-01

    The thermoluminescence of carbonate host rock in the vicinity of known lead-zinc and lead-zinc-fluorite mineralization was reexamined for possible development as an exploration technique. The measurements were made with equipment for determining the thermoluminescence spectrum at closely spaced temperature intervals. Radiation-induced thermoluminescence was also measured. Samples were studied from five localities in Mexico, Southwest Africa, and the United States. Four thermoluminescence properties were found to vary with ''distance-from-ore'' in a systematic manner. These include the glow peak intensity and temperature and the emission spectrum peak energy and full width at half-maximum. For example, in both limestone and dolomite, the high-temperature glow peak intensities are low or negligible within the ore and as the distance from the contact increases the intensity rises rapidly to a maximum, or maxima, and then decreases irregularly to constant value slightly above that in the ore. Depending on the thickness of the ore, the thermoluminescence characteristics associated with the mineralization extended from ten to a hundred or so meters from the ore host rock contact. 5 figures

  15. Fourth intercomparison of personal dosemeters used in US Department of Energy accelerator facilities

    CERN Document Server

    Stewart, R D; Otto, T; Loesch, R M

    2000-01-01

    Personal neutron dosemeters from seven US Department of Energy (DOE) laboratories were mailed to the European Laboratory for Particle Physics (CERN) and irradiated using the well-characterised CERN reference radiation facility (CERF). Neutron dose equivalents determined using the DOE personal dosemeters have been compared to the reference dose equivalent as determined using a tissue-equivalent proportional counter (TEPC). In the 0.5 to 5 mSv dose equivalent range, the comparison of results suggests that the neutron personal dosemeters in use at DOE facilities are capable of estimating dose equivalents for high energy neutrons to within a factor of at least 2 or 3. If a field-specific calibration factor is used to correct the dose equivalent responses, the agreement with the reference dose equivalent for these dosemeters can be improved to better than about 25 to 65at is decoupled from the core in /sup 183,185/Au, becomes the 3/2[532] state (h9/2 parentage) strongly coupled in the doubly-odd /sup 184/Au nucleu...

  16. Investigation of thermoluminescence properties of mobile phone screen displays as dosimeters for accidental dosimetry

    International Nuclear Information System (INIS)

    Mrozik, Anna; Marczewska, B.; Bilski, P.; Kłosowski, M.

    2014-01-01

    The rapid assessment of the radiation dose after unexpected exposure is a task of accidental dosimetry. In case of a radiological accident glasses originating from mobile phone screens, placed usually near the human body, could be used as emergency thermoluminescent (TL) personal dosimeters. The time between irradiation and TL readout is crucial and therefore preparation of the mobile phone screens and their readout conditions should be optimized. The influence of the samples etching, bleaching and selection of the optical filters based on measurement of the emission spectrum of irradiated glass samples during heating for different types of mobile phones were the subjects of our investigation. Obtained results showed that glasses extracted from different brands of mobile phones have different dosimetric properties but all of them give a luminescent signal which can be used to calculate the dose. - Highlights: • Application of acids mixture for more effective glass samples etching. • Bleaching can be performed not only by blue but also by UV light. • The readout with BG39 optical filter improves the TL signal intensity. • The glass samples from touch mobile phones are worse emergency TL dosimeters

  17. Calibration procedures for mammography dosemeters in Poland

    International Nuclear Information System (INIS)

    Gwiazdowska, B.; Ulkowski, P.; Tolwinski, J.; Bulski, W.

    2002-01-01

    Breast cancer is the most frequent tumour in women and the effectiveness of the treatment depends dramatically on the early detection of the disease. That is the reason why in Poland the mammography control examinations are strongly supported by the Centre of Oncology. In Poland there are over 400 mammography units which account for about 300,000 examinations per year. An investigation performed by the Medical Physics Department of the Centre of Oncology in Warsaw at about 100 mammography facilities proved that in most cases the doses absorbed by the patients could be reduced without decrease of image quality. This is one of the reasons why the Polish Secondary Standard Dosimetry Laboratory (SSDL) dealing mainly with calibration of radiotherapy dosemeters is extending its activities and therefore new facilities and equipment adapted for calibration of mammographic dosemeters have been installed. The mammography dosimetry calibration equipment is permanently installed in the same laboratory room where the radiotherapy dosemeters are calibrated. A base of a mammography unit no longer in clinical use, together with its movable system has been adapted to handle ionization chamber holders. An X-ray tube with a 50 kV high frequency generator was also installed. The tube, a Varian type OEG-50-2, (designed for laboratory applications) with molybdenum anode of an anode angle 23,7 deg. and with a large focus, effective size approximately 5 mm 2 , has an inherent filtration of 0,25 mm beryllium. It is installed in a housing with 2mm lead shielding; a cone shaped beam is formed by a system of three collimators

  18. Fundamental study on the characteristics of a radiophotoluminescence glass dosemeter with no energy compensation filter for measuring patient entrance doses in cardiac interventional procedures.

    Science.gov (United States)

    Kato, Mamoru; Chida, Koichi; Moritake, Takashi; Koguchi, Yasuhiro; Sato, Tadaya; Oosaka, Hajime; Tosa, Tetsuo; Kadowaki, Ken

    2014-12-01

    Cardiac interventional procedures have been increasing year by year. However, radiation skin injuries have been still reported. There is a necessity to measure the patient entrance skin dose (ESD), but an accurate dose measurement method has not been established. To measure the ESD, a lot of radiophotoluminescence dosemeters (RPLDs) provide an accurate measurement of the direct actual ESD at the points they are arrayed. The purpose of this study was to examine the characteristics of RPLD to measure the ESD. As a result, X-ray permeable RPLD (with no tin filter) did not interfere with the percutaneous coronary intervention procedure. The RPLD also had good fundamental performance characteristics. Although the RPLD had a little energy dependence, it showed excellent dose and dose-rate linearity, and good angular dependence. In conclusion, by calibrating the energy dependence, RPLDs are useful dosemeter to measure the ESD in cardiac intervention. © The Author 2013. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  19. Basic requirements of dosemeter systems for individual monitoring of external radiation

    International Nuclear Information System (INIS)

    Boehm, J.; Ambrosi, P.

    1985-01-01

    A plea is made for detailed detector independent requirements for dosemeter systems for individual monitoring of external radiation. These requirements should have their origin in the fundamental aspects and concepts of radiation protection for workers, and should be something like a translation of the general principles of individual monitoring into a language easily understandable by producers and users. This work comprises a summary of the general objectives of individual monitoring and discussion of some relevant requirements for dosemeter systems. (orig.) [de

  20. The role of phantom parameters on the response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter

    International Nuclear Information System (INIS)

    Sohrabi, M.; Katouzi, M.

    1992-01-01

    The response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter (NANPD) which can also be used for other albedo dosemeter types was determined on 18 different phantom configurations. The effects of type, geometry, material, thickness, dosemeter-to-phantom angle in particular with the presence of legs were investigated using a Pu-Be neutron source. It was concluded that the slab phantoms (single or double) and circular and elliptical cylinder phantoms seemed to provide a better response, whereas the ICRU sphere geometry does not seem to be appropriate for the calibration of albedo dosemeters. It is interesting to note that the presence of legs maintains the constancy of the response in a situation when a radiation worker bends down during work. (author)

  1. Energy dependence of ulrathin LiF-dosemeters for high energy electrons and high energy X-radiation

    International Nuclear Information System (INIS)

    Kupfer, T.

    1977-02-01

    The energy dependence of ultrathin LiF-dosemeters for high energy electrons (5-40 MeV) and high energy X-radiation (6 MV, 42 MV) is experimentally determined. The experimental values are compared to values calculted earlier by other authors. The influence of the thickness of the dosemeters have been considered by comparison of experimental values for 0.03 mm thick dosemeters and theoretical values for 0.13 mm and 0.38 mm thick ones. Also different commersially available dosemeters have been compared by experiments. It is difficult to draw any other conclutions about the energy dependence than that the variation of the relative responce is within +- 3 percent (2S). However the results seems to be sulficient for clinical applications

  2. Feasibility study of radiophotoluminescent glass rod dosimeter postal dose intercomparison for high energy photon beam

    International Nuclear Information System (INIS)

    Rah, Jeong-Eun; Kim, Siyong; Cheong, Kwang-Ho; Lee, Jeong-Woo; Chung, Jin-Beom; Shin, Dong-Oh; Suh, Tae-Suk

    2009-01-01

    A radiophotoluminescent glass rod dosimeter (GRD) system has recently become commercially available. In this study we evaluated whether the GRD would be suitable for external dosimetric audit program in radiotherapy. For this purpose, we introduced a methodology of the absorbed dose determination with the GRD by establishing calibration coefficient and various correction factors (non-linearity dose response, fading, energy dependence and angular dependence). A feasibility test of the GRD postal dose intercomparison was also performed for eight high photon beams by considering four radiotherapy centers in Korea. In the accuracy evaluation of the GRD dosimetry established in this study, we obtained within 1.5% agreements with the ionization chamber dosimetry for the 60 Co beam. It was also observed that, in the feasibility study, all the relative deviations were smaller than 3%. Based on these results, we believe that the new GRD system has considerable potential to be used for a postal dose audit program

  3. Use of glasses as radiation detectors for high doses

    International Nuclear Information System (INIS)

    Caldas, L.

    1989-08-01

    Glass samples were tested in relation to the possibility of use in high dose dosimetry in medical and industrial areas. The main characteristics were determined: detection threshold, reproducibility, response to gamma radiation of 137 Cs and 6 Co and thermal decay at ambient temperature, with the use of optical absorption and thermoluminesce techniques. (author) [pt

  4. Coincidence in the dose estimation in a OEP by different methods

    International Nuclear Information System (INIS)

    Guerrero C, C.; Arceo M, C.; Brena V, M.

    2007-01-01

    The case of an apparent overexposure to radiation according to that indicated for the thermoluminescent dosemeter 81.59 mSv (TLD) of a occupationally exposed hard-working (POE), for that was practiced the study of biological dosimetry. The estimated dose was 0.12 Gy with which was proven the marked dose registration by the TLD dosemeter. It was concluded that both doses are the same ones. (Author)

  5. Coincidence in the dose estimation in a OEP by different methods; Coincidencia en la estimacion de dosis en un POE por diferentes metodos

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero C, C.; Arceo M, C.; Brena V, M. [ININ, Km. 36.5, Carretera Mexico-Toluca, 52750 La Marquesa, Estado de Mexico (Mexico)]. e-mail: cgc@nuclear.inin.mx

    2007-07-01

    The case of an apparent overexposure to radiation according to that indicated for the thermoluminescent dosemeter 81.59 mSv (TLD) of a occupationally exposed hard-working (POE), for that was practiced the study of biological dosimetry. The estimated dose was 0.12 Gy with which was proven the marked dose registration by the TLD dosemeter. It was concluded that both doses are the same ones. (Author)

  6. Zirconium based bulk metallic glasses

    International Nuclear Information System (INIS)

    Dey, G.K.; Neogy, S.; Savalia, R.T.; Tewari, R.; Srivastava, D.; Banerjee, S.

    2006-01-01

    Metallic glasses have come into prominence in recent times because their nanocrystalline atomic arrangement imparts many useful and unusual properties to these metallic solids. In this study, bulk glasses have been obtained in Zr based multicomponent alloy by induction melting these alloys in silica crucibles and casting these in form of rods 3 and 6 mm in diameter in a copper mould

  7. Development of radiophotometric dosemeters with high sensitivity using plastic scintillators as a light intensifier

    International Nuclear Information System (INIS)

    Mesquita, C.H. de; Hamada, M.M.

    1987-01-01

    Rectangular plates of plastic scintillators are developed and their effect as light converter evaluated, when used as film-holder in conventional photography dosemeters. In this dosemeter, the radiation that not interacts in the photographic film can be detected by light photons generation in the plastic scintillators, sensitizing the film. (C.G.C.) [pt

  8. Mensuration of equivalent dose with personal dosemeters and instruments of radiological protection in the new operative magnitudes ICRU, for external fields of beta radiation. Part I. Study of the homogeneity of the response personal dosemeters leaves (cards G-1, TLD-100), in radiation fields of Co{sub 60}; Medicion de dosis equivalente con dosimetros personales e instrumentos de proteccion radiologica en las nuevas magnitudes operativas ICRU, para campos de radiacion beta externos. Parte I. Estudio de la homogeneidad de la respuesta dosimetros personales (tarjetas G-1, TLD-100), en campos de radiacion de Co{sub 60}

    Energy Technology Data Exchange (ETDEWEB)

    Alvarez R, J.T. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    1994-01-15

    and without personal dosemeters, and for the bottom readings and sensibility of the reader equipment; and of the adjustment kindness to the normal for each game of data.The main conclusions that they were derived of this study they are: i.That the reader equipment presents differences statistically significant ({alpha}=5%) in their sensibilities; that which is necessary correct the readings of the dosemeter, making stand out the necessity of building the curves of operation of the equipment for the bottom and the sensibility. ii. The filter of A1 is not equivalent among the different personal dosemeters i i.For this sample of 40 cards with personal dosemeters to the one fewer exist: eleven populations different statistic significant, ({alpha}) to the response glass 1 five for the response of the glass 2, respectively. i v. That exist difference significant statistical, ({alpha} =5%) among the repetitions; what implies that it is necessary to control but strictly the experimental parameters to the one moment of carrying out the irradiations, (field size, correction for time of obturator of the source, distances source-detecting, etc), since the influence of the reader equipment it has already been considered. (Author)

  9. Thermoluminescent dosimetry of beta radiations of {sup 90} Sr/ {sup 90} Y using amorphous ZrO{sub 2}; Dosimetria termoluminiscente de radiaciones beta de {sup 90} Sr/ {sup 90} Y usando ZrO{sub 2} amorfo

    Energy Technology Data Exchange (ETDEWEB)

    Rivera M, T. [CICATA-Legaria, IPN, Legaria Num. 694, 11500 Mexico D.F. (Mexico); Olvera T, L.; Azorin N, J.; Barrera R, M.; Soto E, A.M. [UAM-I, 09340 Mexico D.F. (Mexico)

    2005-07-01

    In this work the results of studying the thermoluminescent properties (Tl) of the zirconium oxide in its amorphous state (ZrO{sub 2}-a) before beta radiations of {sup 90} Sr/ {sup 90} Y are presented. The amorphous powders of the zirconium oxide were synthesized by means of the sol-gel technique. The sol-gel process using alkoxides like precursors, is an efficient method to prepare a matrix of zirconium oxide by hydrolysis - condensation of the precursor to form chains of Zr-H{sub 3} and Zr-O{sub 2}. One of the advantages of this technique is the obtention of gels at low temperatures with very high purity and homogeneity. The powders were characterized by means of thermal analysis and by X-ray diffraction. The powders of ZrO{sub 2}-a, previously irradiated with beta particles of {sup 90} Sr/{sup 90} Y, presented a thermoluminescent curve with two peaks at 150 and 257 C. The dissipation of the information of the one ZrO{sub 2}-a was of 40% the first 2 hours remaining constant the information for the following 30 days. The reproducibility of the information was of {+-} 2.5% in standard deviation. The studied characteristics allow to propose to the amorphous zirconium oxide as thermoluminescent dosemeter for the detection of beta radiation. (Author)

  10. Thermoluminescent dosimetry of beta radiations of {sup 90} Sr/ {sup 90} Y using ZrO{sub 2}: Eu; Dosimetria termoluminiscente de radiaciones beta de {sup 90} Sr/ {sup 90} Y usando ZrO{sub 2}: Eu

    Energy Technology Data Exchange (ETDEWEB)

    Olvera T, L.; Azorin N, J.; Barrera S, M.; Soto E, A.M. [UAM-I, 09340 Mexico D.F. (Mexico); Rivera M, T. [CICATA-IPN, Legaria 694, 11500 Mexico D.F. (Mexico)

    2005-07-01

    In this work the results of studying the thermoluminescent properties (TL) of the doped zirconium oxide with europium (ZrO{sub 2}: Eu{sup 3+}) before beta radiations of {sup 90}Sr/ {sup 90}Y are presented. The powders of ZrO{sub 2}: Eu{sup 3+} were obtained by means of the sol-gel technique and they were characterized by means of thermal analysis and by X-ray diffraction. The powders of ZrO{sub 2}: Eu{sup 3+}, previously irradiated with beta particles of {sup 90}Sr/ {sup 90}Y, presented a thermoluminescent curve with two peaks at 204 and 292 C respectively. The TL response of the ZrO{sub 2}: Eu{sup 3+} as function of the absorbed dose was lineal from 2 Gy up to 90 Gy. The fading of the information of the ZrO{sub 2}: Eu{sup 3+} was of 10% the first 2 hours remaining almost constant the information by the following 30 days. The ZrO{sub 2} doped with the (Eu{sup 3+}) ion it was found more sensitive to the beta radiation that the one of zirconium oxide without doping (ZrO{sub 2}) obtained by the same method. Those studied characteristics allow to propose to the doped zirconium oxide with europium like thermoluminescent dosemeter for the detection of the beta radiation. (Author)

  11. Investigation of thermoluminescent response of K2YF5:Dy3+ crystals for gamma and X radiation fields

    International Nuclear Information System (INIS)

    Silva, E.C.; Nogueira, M.S.; Faria, L.O.; Khaidukov, N.M.

    2005-01-01

    K 2 YF 5 crystals doped with rare earths have been synthesized with 0 to 100% of Dy 3+ ions optically active ions and an investigation was conducted to test its thermoluminescent (TL) response due to function of Dy 3+ concentration and their response in energy. After being irradiated with gamma and X-rays, it was observed that crystals doped with 1.0% of Dy 3+ feature the best response TL. The main dosimetric peak can be decomposed into three secondary TL peaks, centered in 96.4, 104.9 and 130.7 deg C, respectively, showing a good linearity and reproducibility of the dose measurements. The sensitization process seems to improve response TL and TL peak sensitivity increase to 130, 7 deg C at the expense of TL peak to 104, 9 deg C. The linear coefficient sign TL for K 2 Y 0.09 Dy 0.01 F5 is comparable to that of the dosemeter CaSO 4 : Mn, irradiated with gamma radiation source ( 137 Cs) under the same conditions. Energy dependence measurements show that the answer for X-rays with energy of 41.1 keV is more than 30 times the response to Cs-137, when exposed to the same dose. Due to the main peak in low temperature and the TL high reply to low energy fields, the results reported indicate that the K 2 YF 5 crystals doped with Dy 3+ present great potential for radiation dosimetry in X-rays therapy, clinical dosimetry and also for applications in digital thermoluminescent images

  12. Personnel radiation monitoring by thermoluminescence dosimetry (2000-2001)

    International Nuclear Information System (INIS)

    Mi-Cho-Cho, Daw; Hla-Hla-Win, Daw; Thin-Thin-Kraing

    2001-01-01

    Thermoluminescence dosimetry service was introduced in 1991. Personnel who exposed directly or indirectly to radiation is monitored by thermoluminescent dosimeter. TL materials used for thermoluminescent dosimeter are in the form of thin disc. Personnel whole-body and extremity doses are measured by thermoluminescence dosimetry. The Harshaw Model 4500 TLD reader and Vinten 654E TLD reader are used for TLD evaluation. At present about 600 radiation workers are provided with TLD for routine monitoring. It was found that most personnel had received within permissible dose recommended by the International Commission on Radiological Protection (ICRP). (author)

  13. Deriving staff and public doses in a PET/CT facility from measured radiation levels using thermoluminescent dosimetry

    International Nuclear Information System (INIS)

    Al-Haj, A. N.; Lobriguito, A. M.; Arafah, A.; Parker, R.

    2011-01-01

    The introduction of PET/CT at King Faisal Specialist Hospital and Research Centre for whole body and brain imaging has become favourable for diagnosis of cancer. There is no data available on the PET/CT dose to staff and members of the public for different activities of 18 F [fluorodeoxyglucose (FDG)] and for longer patient holding time. The study aims to determine and evaluate staff and public doses by using thermoluminescent dosemeters monthly readings for a 7-month deployment period and by using direct measurements of dose rates at 30 cm and 1 m distances from the patients after injection. The whole body doses per procedure and per administered activity of 18 F (FDG) were estimated. A dose map inside the PET/CT was generated to provide information of the dose levels in different locations. The Pearson correlation showed a strong correlation (r 2 = 0.71) between the dose per activity and the number of patients. Optimisation of radiation protection of staff and members of the public was investigated and recommendations were given. (authors)

  14. Sensitivity of various thermoluminescent, radiophotoluminescent and photographic detectors to neutrons emitted by a 252Cf source

    International Nuclear Information System (INIS)

    Spurny, Frantisek; Medioni, Roger; Chapuis, A.; Portal, Guy.

    1975-07-01

    The specific sensitivity of various thermoluminescent, radiophotoluminescent and photographic detectors to the neutron spectrum of a 252 Cf source was measured and the effect of the size and composition of the containers in which they might be put was investigated. PB33 radiophotoluminescent glasses, radiothermoluminescent alumina and calcium sulfate were less sensitive to fission neutrons whereas photographic emulsions were more sensitive. The former should be used for γ detection in mixed fields of photons and fission neutrons [fr

  15. Exploratory experimental investigations on post-tensioned structural glass beams

    DEFF Research Database (Denmark)

    Louter, C.; Nielsen, Jens Henrik; Belis, J.

    2013-01-01

    This paper discusses two projects on post-tensioned glass beams, performed at EPFL and DTU, respectively. In these projects small scale glass beams (length of 1.5m and 1m) are post-tensioned by means of steel threaded rods tensioned at the beam ends. The purpose of post-tensioning glass beams...

  16. The thermoluminescence as tool in the reconstruction of volcanic events; La termoluminiscencia como herramienta en la reconstruccion de eventos volcanicos

    Energy Technology Data Exchange (ETDEWEB)

    Ramirez L, A.; Schaaf, P.; Martin del Pozzo, A.L.; Gonzalez M, P. [Instituto de Geofisica, UNAM, C.P. 04500, Mexico D.F. (Mexico)

    2000-07-01

    Within the Mexican land a great number of volcanoes are situated which a considerable part of them are still active. The relevance of dating pomex deposits, ash or lava of these poly genetic volcanoes is to determine the periodicity and magnitude of the volcanic events happened. In this work is presented the preliminary result of the dating by thermoluminescence in a pomex of a pyroclastic flux coming from a volcano in the state of Puebla with the purpose of providing elements to the knowledge which describe the eruptive history of the explosive volcanism at center of Mexico. For the sample dating the volcanic glasses of pomex were separated and it was applied the fine grain technique with a grain size between 4-11 {mu} m. In order to calculate the rate of annual dose it was carried out the following: in the determination of {sup 238} U and {sup 232} Th radioisotope concentration was used the neutron activation technique in a nuclear reactor, in the determination of the K 40 radioisotope was used a scanning electron microscope, the rate of environmental and cosmic dose was measured arranging Tl dosemeters of CaSO{sub 4}: Dy in the sampling place. In order to calculate the paleodoses it was carried out the following: the equivalent dose (Q) was determined starting form the additive method and the supra linearity factor (I) starting from regenerative method and in both methods the irradiated process was realized with a {sup 90} Sr beta source. With the above determinations it was calculated a paleodoses of 231 Gy and a rate of annual dose of 6.074 x 10{sup -3} Gy/year, estimating an age of: Age{sub pomez} = 231 Gy / 6.074 Gy x 10{sup -3} Gy /year = 38030 {+-} 4000 years. (Author)

  17. Dosimetric characteristics of a TLD dosemeter with extremities; Caracteristicas dosimetricas de un dosimetro TLD de extremidades

    Energy Technology Data Exchange (ETDEWEB)

    Molina P, D.; Diaz B, E.; Lien V, R. [Centro de Proteccion e Higiene de las Radiaciones, CPHR, Apdo.Postal 6195, Habana 6, CP 10600, Ciudad Habana (Cuba)

    1999-07-01

    It was designed a TLD dosemeter for the monitoring of the extremities. This one consists in a metallic ring with a circular orifice where is arranged a T L detector of LiF: Mg,Ti (Model JR1152C) 5 x 5 x 0.8 mm{sup 3} covered by a polyethylene fine layer. In this work were studied the dosimetric properties of the dosemeter for its application in the dosimetry of extremities for photonic radiation. the results obtained allow conclude that the designed dosemeter can be used for the extremities monitoring. (Author)

  18. Reactor Gamma Heat Measurements with Calorimeters and Thermoluminescence Dosimeters

    DEFF Research Database (Denmark)

    Haack, Karsten; Majborn, Benny

    1973-01-01

    Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than calorimet......Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than...... calorimeters, but possess advantages such as a small probe size and the possibility of making simultaneous measurements at many different positions. Hence, thermoluminescence dosimeters may constitute a valuable supplement to calorimeters for reactor γ-ray heating measurements....

  19. Performance testing of personal dosemeters from eleven dosimetry services in Sweden

    International Nuclear Information System (INIS)

    Lund, E.; Kylloenen, J.-E.; Grindborg, J.-E.; Lindborg, L.

    2001-01-01

    The Swedish regulation, SSI FS 98:5, requires that radiological workers of category A use dosemeters from an approved personal dosimetry service. The 11 services operating in Sweden at the moment use five different types of dosemeter. All have been tested for their ability to determine H p (10) and some of them to determine H p (0.07) according to the European Commission report Radiation Protection 73, EUR 14852, of 1994. The five unique systems have been tested regarding the angular and energy dependence of the response of the dosemeters. The test points for the determination of H p (10) are all, except one, within the trumpet curve and for the unique systems it is shown that the uncertainty related to angular response at three different energies is within the required ±40% except for the lowest X ray quality 40 kV. The energy dependence dominates over the directional dependence and the choice of radiation quality for calibration is of great importance for the system performance. (author)

  20. New method of preparation of CaSO{sub 4}: Dy for ionizing radiation dosimetry; Nuevo metodo de preparacion del CaSO{sub 4}: Dy para dosimetria de radiacion ionizante

    Energy Technology Data Exchange (ETDEWEB)

    Roman L, J.; Rivera M, T.; Alarcon F, G.; Guzman M, J. [CICATA-IPN, 11500 Mexico D.F. (Mexico); Azorin N, J.; Sosa F, R. [UAM-I, 09340 Mexico D.F. (Mexico); Serrano F, A.K. [Hospital Juarez General de Mexico, 07760 Mexico D.F. (Mexico)]. e-mail: holand_jeos@hotmail.com

    2008-07-01

    In this work some results of the characterization of solid state materials exposed at a gamma radiation beam coming from a {sup 60} Co bomb for medical use are presented. These thermoluminescent dosemeters are prepared and proposed for the ionizing radiations dosimetry using the thermoluminescence method. The passive dosemeters of CaSO{sub 4}: Dy are developed in the polycrystalline powder form being used a new synthesis route by means of the precipitation method. To determine the sensibility of the pellets of CaSO{sub 4}: Dy before gamma radiation, these were exposed before a gamma radiation beam coming from a {sup 60} Co bomb. The morphological and structural characteristics were also studied and present two in this work. (Author)

  1. Validation of a MOSFET dosemeter system for determining the absorbed and effective radiation doses in diagnostic radiology.

    Science.gov (United States)

    Manninen, A-L; Kotiaho, A; Nikkinen, J; Nieminen, M T

    2015-04-01

    This study aimed to validate a MOSFET dosemeter system for determining absorbed and effective doses (EDs) in the dose and energy range used in diagnostic radiology. Energy dependence, dose linearity and repeatability of the dosemeter were examined. The absorbed doses (ADs) were compared at anterior-posterior projection and the EDs were determined at posterior-anterior, anterior-posterior and lateral projections of thoracic imaging using an anthropomorphic phantom. The radiation exposures were made using digital radiography systems. This study revealed that the MOSFET system with high sensitivity bias supply set-up is sufficiently accurate for AD and ED determination. The dosemeter is recommended to be calibrated for energies 80 kVp. The entrance skin dose level should be at least 5 mGy to minimise the deviation of the individual dosemeter dose. For ED determination, dosemeters should be implanted perpendicular to the surface of the phantom to prevent the angular dependence error. © The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  2. Dose estimation by simultaneous measurement of the radon/thoron concentration and the equilibrium factors in air using a passive dosemeter

    International Nuclear Information System (INIS)

    Urban, M.

    1984-03-01

    Responsible for an increased radiation exposure is the inhalation of radon and its short lived daughters. A time integrating passive dosemeter was developed to determine the concentrations of the radon isotopes as well as their equilibrium factors. The α energy spectrum inside a dosemeter is measured by means of a nuclear track detector. The concentrations in air and the equilibrium factors are calculated by using a new mathematical dosemeter model. A small pilot study in houses was done to test the dosemeter. (orig.) [de

  3. Individual TL detector characteristics in automated processing of personnel dosemeters: correction factors as extension to identity codes of dosemeter cards

    International Nuclear Information System (INIS)

    Toivonen, Matti.

    1979-07-01

    One, two and three-component dosemeter cards and their associated processing equipment were developed for personnel monitoring. A novel feature of the TLD system is that the individual sensitivity correction factors of TL detectors for β/γ radiation dosimetry and special timing factors for the readout of neutron detectors are stored on dosemeter cards as an extension of the identity codes. These data are utilized in the automatic TL reading process with the aim of cancelling out the influence of the individual detector characteristics on the measuring results. Stimulation of TL is done with hot nitrogen without removing the detectors from their cards and without any metal contact. Changes in detector characteristics are thus improbable. The reading process can be adjusted in a variety of ways. For example, each detector in the same card can be processed with optimal heating and the specific 250 deg C glow peak of neutron radiation can be roughly separated from the main LiF glow peaks. (author)

  4. Thermoluminescence fast neutron dosimetry by laser heating

    International Nuclear Information System (INIS)

    Mathur, V.K.; Brown, M.D.; Braeunlich, P.

    1984-01-01

    Heating rates in excess of 10 4 K.sec -1 have been achieved for thin layers of TL dosemeters by laser heating. The high heating rate improves the signal to noise ratio up to a factor of 10 3 . Thus sensitive thin film fast neutron dosemeters with negligible self-shielding have become a practical reality. Thin samples of CaSO 4 :Dy have been investigated for their response to fast neutrons from a Pu-Be source and a 14.6 MeV neutron generator by using a hydrogenous radiator. A 15 watt CO 2 laser was focussed on the thin TLD layer to a spot size of less than 1 mm to heat it. An exposure of a few tens of milliseconds was sufficient to obtain a TLD curve, which was displayed and processed by a wave form digitiser. The laser spot could be scanned over the TLD sample by a x-y positioner and a large number of observations were obtained on each sample. Preliminary results show that it is possible to obtain a figure of merit of approx. 5% in a mixed n, γ field. A practical design for a fast neutron dosemeter is proposed. (author)

  5. Performance of a PADC personal neutron dosemeter at simulated and real workplace fields of the nuclear industry

    International Nuclear Information System (INIS)

    Fiechtner, A.; Boschung, M.; Wernli, C.

    2007-01-01

    In the framework of the EVIDOS (Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields) project, funded by the EC, measurements with PADC personal neutron dosemeters were carried out at several workplace fields of the nuclear industry and at simulated workplace fields. The measured personal neutron dose equivalents of the PADC personal neutron dosemeter are compared with values that were assessed within the EVIDOS project by other partners. The detection limits for different spectra types are given. In cases were the neutron dose was too low to be measured by the PADC personal neutron dosemeter, the response is estimated by convoluting the responses to monoenergetic neutrons with the dose energy distribution measured within EVIDOS. The advantages and limitations of the PADC personal neutron dosemeter are discussed. (authors)

  6. Measuring the absorbed dose in critical organs during low rate dose brachytherapy with 137 Cs using thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Torres, A.; Gonzalez, P.R.; Furetta, C.; Azorin, J.; Andres, U.; Mendez, G.

    2003-01-01

    Intracavitary Brachytherapy is one of the most used methods for the treatment of the cervical-uterine cancer. This treatment consists in the insertion of low rate dose 137 Cs sources into the patient. The most used system for the treatment dose planning is that of Manchester. This planning is based on sources, which are considered fixed during the treatment. However, the experience has shown that, during the treatment, the sources could be displaced from its initial position, changing the dose from that previously prescribed. For this reason, it is necessary to make measurements of the absorbed dose to the surrounding organs (mainly bladder and rectum). This paper presents the results of measuring the absorbed dose using home-made LiF: Mg, Cu, P + Ptfe thermoluminescent dosimeters (TLD). Measurements were carried out in-vivo during 20 minutes at the beginning and at the end of the treatments. Results showed that the absorbed dose to the critical organs vary significantly due to the movement of the patient during the treatment. (Author)

  7. Improvement of personal dosimetry - Digital pocket dosemeter

    International Nuclear Information System (INIS)

    Radalj, Z.; Cerovac, Z.; Cerovac, H.; Brumen, V.; Prlic, I.

    1996-01-01

    Physical dosimetric surveillance of professional groups working with various radiation sources is a regular procedure in Croatia, established almost 40 years ago. Data available point out that the majority of professionals under surveillance are those employed in medical facilities, most of them working with X-ray sources. Depending on the nature of professional activities, personnel occupationally exposed to radiation sources are obliged to wear either film badge, TLD or film+TLD badge. Unfortunately, due to the line of data processing, all standard dosemeters have the same disadvantage i.e. up to 40 days delay in dose reporting, regarding the time of actual exposure. The significance of such a delay raises in cases when radiation dose was received within the short time or when technical failure on the operating unit(s) is suspected. Bearing this in mind, the additional dosimetric monitoring becomes an imperative. Therefore, we decided to introduce a palette of digital pocket dosemeters, meant to be used in different workplaces in the radiation zone, each of them being adjusted to the specificities of a particular workplace

  8. A neutron dosemeter for nuclear criticality accidents.

    Science.gov (United States)

    d'Errico, F; Curzio, G; Ciolini, R; Del Gratta, A; Nath, R

    2004-01-01

    A neutron dosemeter which offers instant read-out has been developed for nuclear criticality accidents. The system is based on gels containing emulsions of superheated dichlorodifluoromethane droplets, which vaporise into bubbles upon neutron irradiation. The expansion of these bubbles displaces an equivalent volume of gel into a graduated pipette, providing an immediate measure of the dose. Instant read-out is achieved using an array of transmissive optical sensors which consist of coupled LED emitters and phototransistor receivers. When the gel displaced in the pipette crosses the sensing region of the photomicrosensors, it generates a signal collected on a computer through a dedicated acquisition board. The performance of the device was tested during the 2002 International Accident Dosimetry Intercomparison in Valduc, France. The dosemeter was able to follow the initial dose gradient of a simulated accident, providing accurate values of neutron kerma; however, the emulsion was rapidly depleted of all its drops. A model of the depletion effects was developed and it indicates that an adequate dynamic range of the dose response can be achieved by using emulsions of smaller droplets.

  9. A neutron dosemeter for nuclear criticality accidents

    International Nuclear Information System (INIS)

    D'Errico, F.; Curzio, G.; Ciolini, R.; Del Gratta, A.; Nath, R.

    2004-01-01

    A neutron dosemeter which offers instant read-out has been developed for nuclear criticality accidents. The system is based on gels containing emulsions of superheated dichlorodifluoromethane droplets, which vaporise into bubbles upon neutron irradiation. The expansion of these bubbles displaces an equivalent volume of gel into a graduated pipette, providing an immediate measure of the dose. Instant read-out is achieved using an array of transmissive optical sensors which consist of coupled LED emitters and phototransistor receivers. When the gel displaced in the pipette crosses the sensing region of the photo microsensors, it generates a signal collected on a computer through a dedicated acquisition board. The performance of the device was tested during the 2002 International Accident Dosimetry Intercomparison in Valduc (France)). The dosemeter was able to follow the initial dose gradient of a simulated accident, providing accurate values of neutron kerma; however, the emulsion was rapidly depleted of all its drops. A model of the depletion effects was developed and it indicates that an adequate dynamic range of the dose response can be achieved by using emulsions of smaller droplets. (authors)

  10. Thermoluminescent response of LaAlO3:Pr

    International Nuclear Information System (INIS)

    Morales H, A.; Zarate M, J.; Azorin N, J.; Rivera M, T.

    2015-10-01

    In this study, the thermoluminescence response of doped lanthanum aluminate (LaAlO 3 ) with praseodymium ion (Pr) obtained by the Pechini method and drying by the spraying technique Spry Dryer was studied. The obtained powders were analyzed structurally by the X-ray diffraction technique; the morphological characterization was by the scanning electron microscopy technique. The obtained powders at 800 degrees C presented crystallinity and showed a Rhombohedral crystal structure, this phase was observed by X-ray diffraction patterns. Thermoluminescence response of LaAlO 3 :Pr showed a brightness curve with a peak centered at 157 degrees C. The sensitivity of the doped samples was improved about 90 times in comparison with the undoped sample. Thermoluminescence response in function of the wavelength showed a maximum at 230 nm, reproducibility of thermoluminescence response was ±50%. Also the fading in thermoluminescence response was studied. (Author)

  11. Dosimetric properties of germanium doped calcium borate glass subjected to 6 MV and 10 MV X-ray irradiations

    Science.gov (United States)

    Tengku Kamarul Bahri, T. N. H.; Wagiran, H.; Hussin, R.; Saeed, M. A.; Hossain, I.; Ali, H.

    2014-10-01

    Germanium doped calcium borate glasses are investigated in term of thermoluminescence properties to seek their possibility to use as glass radiation dosimeter. The samples were exposed to 6 MV, and 10 MV photon beams in a dose range of 0.5-4.0 Gy. There is a single and broad thermoluminescence glow curve that exhibits its maximum intensity at about 300 °C. Linear dose response behavior has been found in this dose range for the both photon energies. Effective atomic number, TL sensitivity, and reproducibility have also been studied. It is found that the sensitivity of germanium doped sample at 6 MV is only 1.28% and it is superior to the sensitivity at 10 MV. The reproducibility of germanium doped sample is good with a percentage of relative error less than 10%. The results indicate that this glass has a potential to be used as a radiation dosimetry, especially for application in radiotherapy.

  12. Occupational exposure of nuclear medicine personnel

    International Nuclear Information System (INIS)

    Roessler, M.

    1982-01-01

    The results are given of measurements of the radiation burden of personnel in departments of nuclear medicine in the years 1979 to 1981 using film dosemeters and ring thermoluminescence dosemeters evaluated by the national personnel dosemeter service. The relations are examined of the exposure of hands and the preparation of radiopharmaceuticals and especially their use for examinations. Certain organizational measures are indicated for reducina radiation burden in a laboratory for the preparation of radiopharmaceuticals. The results of measurements and evaluations of radiation burden of personnel of nuclear medicine departments are confronted with conclusions published in the literature. (author)

  13. A new TLD system for space research

    International Nuclear Information System (INIS)

    Feher, I.; Deme, S.; Szabo, B.; Vagvoelgyi, J.; Szabo, P.P.; Csoeke, A.; Ranky, M.; Akatov, Yu.A.

    1980-06-01

    A small, portable, vibration and shock resistant thermoluminescent dosemeter (TLD) system was developed to measure the cosmic radiation dose on board of a spacecraft. The TLD system consists of a special bulb dosemeter and a TLD reader. The measuring dose range of the TLD system is from 10 μGy up to 100 mGy. The TLD reader can operate on a battery; its electrical power consumption is about 5 W, its volume is about 1 dm 3 and its mass is about 1 kg. Details are given of the construction and technical parameters of the dosemeter and reader. (author)

  14. Tl response of LiF: Mg, Cu, P + PTFE (Mexico) and GR200A (China) dosemeters

    International Nuclear Information System (INIS)

    Gonzalez M, P.R.; Azorin N, J.; Furetta, C.

    2003-01-01

    TLD-100 was the commercial dosemeter more known since some decades ago. This dosemeter was considered for many research groups as the reference material for developing new Tl materials for ionizing radiation dosimetry. Actually it seems that TLD-100 is going to be replaced by the Chinese material GR200A, as reference material due that this material in addition to be considered as a tissue equivalent material, is 30 to 35 times more sensitive that TLD- 100. Results of the study of the Tl response of LiF: Mg, Cu, P + Ptfe developed at ININ-Mexico, comparing them with those of GR200A are presented. These results showed that the sensitivity (s = Tl intensity/weight x dose) to gamma radiation of the LiF: Mg, Cu, P + Ptfe dosimeters was 4.34 meanwhile that of the commercial dosemeter was 3.41. Detection threshold of the dosemeters studied was 2.22 and 0.52 μGy respectively Repeatability after 10 cycles Irradiation-reading-annealing was ±1.39% and ±1.86% respectively. Both types of dosemeters presented a linear response as a function of gamma radiation in the range from 0.02 mGy and 100 Gy. (Author)

  15. Official dosimetry with individual electronic dosemeters - the concept in Germany

    International Nuclear Information System (INIS)

    Czarwinski, R.; Kaulard, J.; Pfeffer, W.

    2005-01-01

    Full text: Presently, in Germany passive dosemeters (film batches, RPL, TLD) are used for the official individual monitoring of occupational exposed personal. The application of electronic individual dosemeters (EPD) is carried out mainly for the operative radiation protection control particularly in nuclear power engineering companies, big hospitals und research centres. This means in such institutions double monitoring exists - legally and operatively. A crucial advantage of EPD compared to passive dosemeter is the possibility to adapt the monitoring period to the working time in the control area, e.g. a job related monitoring is available. Germany started a project for an optimized implementation of EPD into the official dosimetry. Objective of the project whose results will be described in the paper is the harmonization of radiation protection of companies workers and outside workers concerning the record of individual and job related exposures. The approval of the EPD by the Federal and Laender authorities as an official dosimetric system demands the implementation of special requirements concerning the data acquisition, data transfer and data evaluation. These issues are focuses of the Federal research project, supervised by the Federal Office for Radiation Protection (BfS) and performed by the Gesellschaft fuer Anlagen- und Reaktorsicherheit (GRS), Cologne to develop a concept in the first stage. (author)

  16. Absorbed dose determination in photon fields using the tandem method

    International Nuclear Information System (INIS)

    Marques Pachas, J.F.

    1999-01-01

    The purpose of this work is to develop an alternative method to determine the absorbed dose and effective energy of photons with unknown spectral distributions. It includes a 'tandem' system that consists of two thermoluminescent dosemeters with different energetic dependence. LiF: Mg, Ti, CaF 2 : Dy thermoluminescent dosemeters and a Harshaw 3500 reading system are employed. Dosemeters are characterized with 90 Sr- 90 Y, calibrated with the energy of 60 Co and irradiated with seven different qualities of x-ray beams, suggested by ANSI No. 13 and ISO 4037. The answers of each type of dosemeter are adjusted to a function that depends on the effective energy of photons. The adjustment is carried out by means of the Rosenbrock minimization algorithm. The mathematical model used for this function includes five parameters and has a gauss and a straight line. Results show that the analytical functions reproduce the experimental data of the answers, with a margin of error of less than 5%. The reason of the answers of the CaF 2 : Dy and LiF: Mg, Ti, according to the energy of the radiation, allows us to establish the effective energy of photons and the absorbed dose, with a margin of error of less than 10% and 20% respectively

  17. Thermoluminescent measurement in space radiation dosimetry

    International Nuclear Information System (INIS)

    Chen Mei; Qi Zhangnian; Li Xianggao; Huang Zengxin; Jia Xianghong; Wang Genliang

    1999-01-01

    The author introduced the space radiation environment and the application of thermoluminescent measurement in space radiation dosimetry. Space ionization radiation is charged particles radiation. Space radiation dosimetry was developed for protecting astronauts against space radiation. Thermoluminescent measurement is an excellent method used in the spaceship cabin. Also the authors mentioned the recent works here

  18. Development of CaSO4:Dy based ring dosemeter for extremity monitoring of radiation workers in India

    International Nuclear Information System (INIS)

    Srivastava, Kshama; Meenal, P.; Bhagat, R.V.; Singh, R.; Kolambe, D.H.; Chougaonkar, M.P.; Sapra, B.K.

    2014-01-01

    Extremity dosemeters are required to be worn in cases where the dose to extremities is expected to be significantly greater than the dose to the whole body. In India, CaSO 4 :Dy based three element personnel monitoring TLD badge worn at chest and wrist level are used for whole body and extremity monitoring, respectively. Presently no official/legal finger dosemeter is available for extremity monitoring for radiation workers in the country. To cater to the long standing requirement of finger dosimeter, a new compact three-element Extremity Ring Badge Dosemeter (ERBD) has been developed for measurement of the equivalent doses received by the extremities of radiation workers in terms of operational quantity Hp(0.07). It was aimed to meet the performance requirement of IEC/ISO standards. This paper gives the design detail and result of experimental studies of ERB dosemeter

  19. Dosimetric evaluation of 252Cf beam for use in radiobiology studies at Hiroshima University

    International Nuclear Information System (INIS)

    Hoshi, M.; Takeoka, S.; Tsujimura, T.; Kuroda, T.; Kawami, M.; Sawada, S.

    1988-01-01

    This report provides reliable tissue kerma in free air data by characterising the 252 Cf irradiating system and by comparing three dosimetry methods, paired chambers (ICRU 1977, 1978), Fricke and thermoluminescence dosemeter. (author)

  20. Systematic processing of the external individual monitoring data obtained by photographic film and capacitor dosemeters

    International Nuclear Information System (INIS)

    Krueger, F.; Shtuk, D.

    1983-01-01

    Individual radiation load measurements have been carried out both with photographic film and capacitor type dosemeters (CD). Systematic statistics methods were applied for the obtained data processing. The CD data were shown to exceed those measured with the film dosemeters. Specific features of the CD application are revealed that should be taken into account in practice [ru

  1. Thermoluminescence from natural calcites

    International Nuclear Information System (INIS)

    Calderon, T.; Jaque, F.; Coy-yll, R.

    1984-01-01

    Thermoluminescence (TL) as well as absorption and EPR spectra of x-irradiated natural calcites have been obtained. Irradiation produces UV absorption bands and a decrease of the Mn 2+ EPR spectrum. A correlation of each TL peak with the bleaching steps of UV absorption bands and the recovering in intensity of the Mn 2+ EPR spectrum has been found. These experimental results support a new model for the radiation damage and thermoluminescence process in calcites. The main point in this model is that holes become trapped at impurities, and the electrons are trapped at dislocations in the form of CO 3 3- . (author)

  2. Mensuration of equivalent dose with personal dosemeters and instruments of radiological protection in the new operative quantities ICRU, for external fields of beta radiation. Part II. I study of the angular response of personal dosemeters TLD-100 in secondary patron fields of beta radiation ({sup 90}Sr / {sup 90}Y); Medicion de dosis equivalente con dosimetros personales e instrumentos de proteccion radiologica en las nuevas magnitudes operativas ICRU, para campos de radiacion beta externos. Parte II. Estudio de la respuesta angular de dosimetros personales TLD-100 en campos patrones secundarios de radiacion beta ({sup 90}Sr/{sup 90}Y)

    Energy Technology Data Exchange (ETDEWEB)

    Alvarez R, J.T. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    1994-01-15

    The objective of this work is to carry out one of the possible ones test type for personal dosemeters TLD, under the recomendations of the ICRU 39, ICRU 43 and the draft of the norm ISO 6980,(1992), with the purpose of verifying the capacity of these detectors to carry out the operative unit: H' (0.07;{alpha}). Since H' (O. 07;{alpha}) this defined one in an expanded field, one of these tests type consist on determining the angular response of these detectors. 20 personal dosemeters TLD-100 was used, (card marks: Harshaw, Model: G-1, with two glasses of TLD-100 absorbed in teflon; the portadosemeters has two windows, a free one and another with a filter of Pb of 171.0 mg cm{sup -2}); these dosemeters they were previously selected, [to see, {sup S}tudy of the Homogeneity of the response of Personal Dosemeters (Cards G-l, TLD-100) in Radiation of Countrysides of {sup 60}Co{sup ,} J.T. Alvarez R. Technician Report GSR/IT/0001/94].The irradiations to effectued in secondary countryside of radiation beta of {sup 90}Sr/{sup 90}Y. The study was undertaken by means of an experimental design of blocks random that contemplate the following variables: intensity of the radiation source, (1850 MBq and 74 MBq); position of irradiation, (four positions); incidence of angle of the radiation (0, 15, 30, 45, 60 and 75 grades) and the absorbed dose in air, (0.005, 0.010, 0.020, 0.050 and 0.100 Gy). Then null hypothesis it was to suppose that there was not difference among the stockings of each treatment, to used the statistical of Duncan to carry out tests of stockings at a level of significance of 5%.These tests of stockings throw the following results in those variables of the experimental design: The irradiations carried out so much with the source pattern secondary of {sup 90}Sr/{sup 90}Y of 1850 MBq and of 74 MBq, they are equivalent reason why they can be used indistinctly. The responses of each one of the glasses of the card are strongly anisotropic for each glass

  3. New generation of ''legal'' dosemeters

    International Nuclear Information System (INIS)

    Fletcher, R.

    1991-01-01

    In the early 1980s research at the National Radiological Protection Board in the UK proved the feasibility of using solid state electronics in an entirely new dosimetry technology - capable of reaching right down into the low energy photon ranges, and able to detect beta radiation. In 1988 the NRPB undertook a joint venture with Siemens Plessey Controls to develop a marketable personal dosemeter meeting full Health and Safety Executive approval as a ''legal'' instrument. The Electronic Personal Dosemeter (EPD) was thus conceived, and will reach the pre-production stage early this year. The EPD makes use of state-of-the-art silicon integrated circuit technology, with a custom amplifier and microprocessor system. The liquid crystal display continuously shows the accumulated short-term penetrating dose in terms of the Hp (10) unit, and can also show superficial dose and dose rates. Because the EPD must be continuously powered, the custom lithium battery was commissioned to ensure a minimum service interval of 12 months. The EPD is the size and weight of a small pocket pager. Although dose data can be read directly from the EPD, a comprehensive data management system is needed to effect real-life use in industry. The EPD thus communicates by infra-red link to a reader unit which interfaces an IBM-compatible PC, allowing authorized personnel to read the dose memories and perform dose alarm threshold settings. (author)

  4. On-field evaluation of operator lens protective devices in interventional radiology

    International Nuclear Information System (INIS)

    Strocchi, S.; Chiaravalli, A.; Veronese, I.; Novario, R.

    2016-01-01

    The recent publication of the Euratom Directive 2013/59, adopting the reduction of eye lens dose limits from 150 to 20 mSv y"-"1, calls for the development of new tools and methodologies for evaluating the eye lens dose absorbed by the medical staff involved in interventional radiology practices. Moreover, the effectiveness of the protective devices, like leaded glasses, which can be employed for radiation protection purposes, must be tested under typical exposure scenarios. In this work, eye lens dose measurements were carried out on an anthropomorphic phantom simulating a physician bound to perform standard interventional neuroradiology angiographic procedures. The correlation between eye lens doses, in terms of Hp(0.07), and the equivalent dose [again in terms of Hp(0.07)] monthly measured with thermoluminescent dosemeters placed above the lead apron at the chest level was studied, in the presence and in the absence of different types of leaded glasses. (authors)

  5. Characteristic of 120 degree C thermoluminescence peak of iceland spar

    International Nuclear Information System (INIS)

    Lu Xinwei; Han Jia

    2006-01-01

    The basic characteristic of 120 degree C thermoluminescence peak of iceland spar was studied. The experimental result indicates the longevity of 120 degree C thermoluminescence peak of iceland spar is about 2 h under 30 degree C. The thermoluminescence peak moves to the high temperature when the heating speed increasing. The intensity of 120 degree C thermoluminescence peak of iceland spar is directly proportional to radiation dose under 15 Gy. (authors)

  6. Unbiased metal oxide semiconductor ionising radiation dosemeter

    International Nuclear Information System (INIS)

    Kumurdjian, N.; Sarrabayrouse, G.J.

    1995-01-01

    To assess the application of MOS devices as low dose rate dosemeters, the sensitivity is the major factor although little studies have been performed on that subject. It is studied here, as well as thermal stability and linearity of the response curve. Other advantages are specified such as large measurable dose range, low cost, small size, possibility of integration. (D.L.)

  7. Numerical analysis of thermoluminescence glow curves

    International Nuclear Information System (INIS)

    Gomez Ros, J. M.; Delgado, A.

    1989-01-01

    This report presents a method for the numerical analysis of complex thermoluminescence glow curves resolving the individual glow peak components. The method employs first order kinetics analytical expressions and is based In a Marquart-Levenberg minimization procedure. A simplified version of this method for thermoluminescence dosimetry (TLD) is also described and specifically developed to operate whit Lithium Fluoride TLD-100. (Author). 36 refs

  8. Dosimetric properties of germanium doped calcium borate glass subjected to 6 MV and 10 MV X-ray irradiations

    Energy Technology Data Exchange (ETDEWEB)

    Tengku Kamarul Bahri, T.N.H., E-mail: tnhidayah2@gmail.com [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 UTM Johor Bahru (Malaysia); Wagiran, H.; Hussin, R.; Saeed, M.A. [Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81310 UTM Johor Bahru (Malaysia); Hossain, I. [Department of Physics, College of Science and Arts, King Abdul Aziz University, 21911 Rabigh (Saudi Arabia); Ali, H. [Department of Radiotherapy and Oncology, Hospital Sultan Ismail, 81100 Johor Bahru (Malaysia)

    2014-10-01

    Highlights: •The TL properties of 29.9CaO–70B{sub 2}O{sub 3}: 0.1GeO{sub 2} glass has been investigated. •We exposed glass samples to 6 MV and 10 MV in a dose range of 0.5–4.0 Gy. •This glass has a potential material to be used for application in radiotherapy. -- Abstract: Germanium doped calcium borate glasses are investigated in term of thermoluminescence properties to seek their possibility to use as glass radiation dosimeter. The samples were exposed to 6 MV, and 10 MV photon beams in a dose range of 0.5–4.0 Gy. There is a single and broad thermoluminescence glow curve that exhibits its maximum intensity at about 300 °C. Linear dose response behavior has been found in this dose range for the both photon energies. Effective atomic number, TL sensitivity, and reproducibility have also been studied. It is found that the sensitivity of germanium doped sample at 6 MV is only 1.28% and it is superior to the sensitivity at 10 MV. The reproducibility of germanium doped sample is good with a percentage of relative error less than 10%. The results indicate that this glass has a potential to be used as a radiation dosimetry, especially for application in radiotherapy.

  9. Dosimetric properties of germanium doped calcium borate glass subjected to 6 MV and 10 MV X-ray irradiations

    International Nuclear Information System (INIS)

    Tengku Kamarul Bahri, T.N.H.; Wagiran, H.; Hussin, R.; Saeed, M.A.; Hossain, I.; Ali, H.

    2014-01-01

    Highlights: •The TL properties of 29.9CaO–70B 2 O 3 : 0.1GeO 2 glass has been investigated. •We exposed glass samples to 6 MV and 10 MV in a dose range of 0.5–4.0 Gy. •This glass has a potential material to be used for application in radiotherapy. -- Abstract: Germanium doped calcium borate glasses are investigated in term of thermoluminescence properties to seek their possibility to use as glass radiation dosimeter. The samples were exposed to 6 MV, and 10 MV photon beams in a dose range of 0.5–4.0 Gy. There is a single and broad thermoluminescence glow curve that exhibits its maximum intensity at about 300 °C. Linear dose response behavior has been found in this dose range for the both photon energies. Effective atomic number, TL sensitivity, and reproducibility have also been studied. It is found that the sensitivity of germanium doped sample at 6 MV is only 1.28% and it is superior to the sensitivity at 10 MV. The reproducibility of germanium doped sample is good with a percentage of relative error less than 10%. The results indicate that this glass has a potential to be used as a radiation dosimetry, especially for application in radiotherapy

  10. Measurements of eye lens doses in interventional cardiology using OSL and electronic dosemeters

    International Nuclear Information System (INIS)

    Sanchez, R.M.; Vano, E.; Fernandez, J.M.; Ginjaume, M.; Duch, M.A.

    2014-01-01

    The purpose of this paper is to test the appropriateness of OSL and electronic dosemeters to estimate eye lens doses at interventional cardiology environment. Using TLD as reference detectors, personal dose equivalent was measured in phantoms and during clinical procedures. For phantom measurements, OSL dose values resulted in an average difference of 215 % vs. TLD. Tests carried out with other electronic dosemeters revealed differences up to ±20 % versus TLD. With dosemeters positioned outside the goggles and when TLD doses were >20 μSv, the average difference OSL vs. TLD was 29 %. Eye lens doses of almost 700 μSv per procedure were measured in two cases out of a sample of 33 measurements in individual clinical procedures, thus showing the risk of high exposure to the lenses of the eye when protection rules are not followed. The differences found between OSL and TLD are acceptable for the purpose and range of doses measured in the survey (authors)

  11. Development of measure methods of radon in indoor air

    International Nuclear Information System (INIS)

    Yaginuma, L.T.; Pela, C.A.; Navas, E.A.; Ghilardi, A.J.P.

    1992-01-01

    The development of some conventional measuring methods, aiming obtain an estimation of radon concentration in air, mainly in indoor air is described, including the charcoal absorption collector, Lucas cell and thermoluminescent dosemeters. (C.G.C)

  12. Two miniature electronic dosemeters Yperwatch - Gamcard

    International Nuclear Information System (INIS)

    Clavel, B.; Jaillon, J.

    1993-01-01

    Yperwatch and Gamcard were miniature electronic gamma personal dose and dose rate monitors. The Yperwatch is mounted in a wristwatch and the Gamcard in a case of the size of a standard credit card. The technology provides users with all the performance of a standard professional electronic dosemeter (integrated dose, dose rate, audible alarms, time in use). Yperwatch and Gamcard are based on well-proven electronic dosimetry technology, but with a greater ease of use. (author)

  13. Evaluations of absorbed dose ratio factor of Al2O3 dosemeter in radiotherapy photon beams using cavity theory

    International Nuclear Information System (INIS)

    Zhu, J.; Chen, S.; Chen, L.; Liu, X.

    2012-01-01

    The aim of the work was to evaluate the absorbed dose ratio factor f md of an Al 2 O 3 dosemeter to water in photon radiotherapy beams using cavity theory. Burlin theory was used for calculating of this ratio. The effective mass attenuation coefficient β was obtained by comparing Monte Carlo simulations in monoenergetic photon beams. The evaluations of the absorbed dose ratio factor f md were studied for Al 2 O 3 dosemeters of different sizes, which were placed at various depths of the water phantom in different radiation field sizes of Mohan's 6, 10 and 15-MV X-rays. Beyond the build-up region, the variation of f md increases by 0.25 % as the depth increases from 4 to 10 cm. The maximum variation due to different dosemeter sizes is 8.3 %. The difference in the f md due to different radiation field sizes is 1.5 %. The effect of the dosemeter size cannot be neglected. The difference in the f md due to the radiation field sizes of different beams would increase as the dosemeter size increases. (authors)

  14. Use of TLD in radiation field studies in different materials

    International Nuclear Information System (INIS)

    Carron, W.; Ghilardi Netto, T.

    1987-01-01

    The dose conversion factor of a thermoluminescent dosemeter of LiF for water, wood and soil is determined. The dose curves in depth relatioship as well as the regression equations are presented. (M.A.C.) [pt

  15. Thermoluminescent dosimetry and assessment of personal dose

    International Nuclear Information System (INIS)

    Boas, J.F.; Martin, L.J.; Young, J.G.

    1982-01-01

    Thermoluminescence is discussed in terms of the energy band structure of a crystalline solid and the trapping of charge carriers by point defects. Some general properties of thermoluminescent materials used for dosimetry are outlined, with thermoluminescence of CaSO 4 :Dy being described in detail. The energy response function and the modification of the energy response of a dosimeter by shielding are discussed. The final section covers the connection between exposure, as recorded by a TLD badge, and the absorbed dose to various organs from gamma radiation in a uranium mine; the conversion from absorbed dose to dose equivalent; and uncertainties in assessment of dose equivalent

  16. The personnel dosimetry record keeping system at AEE Winfrith

    International Nuclear Information System (INIS)

    Gill, D.W.

    1979-09-01

    Since May 1978 the exposure of personnel to external radiation has been assessed by Thermoluminescent Dosimetry, (TLD). The dosemeter consists of a TLD card similar to that used by the National Radiological Protection Board, held in a plastic badge designed at AEE Winfrith, and used in conjunction with a D A Pitman Ltd Type 605 Automatic Reader. The report describes the dosemeter, the operation of the dosimetry service and the system for maintaining a computerised record keeping system. (author)

  17. Characterization of a gamma radiation dosimeter based of glass doped with copper

    International Nuclear Information System (INIS)

    Laameri, Mohamed Hadi; Ben Mansour, Issam

    2010-01-01

    Commercial sodo-calcic silicate glass was studied by thermo luminescence in order to evaluate its potential like material sensitive to the gamma radiation for dosimetric application. We have examined in particular the effect of doping glass copper ion exchange C U-N A for different concentrations technique and multiple conditions of doping on luminescent thermo sensitivity on a very wide range of doses ranging from 10 mGy up to 100 kGy. We have also tried to explain the origin of thermoluminescence observed by exploiting the doped and non-doped samples EPR spectra.

  18. NUCLEAR HEATING IN LIF DOSEMETERS IN A FUSION NEUTRON FIELD, TRIAL OF DIRECT COMPARISON OF EXPERIMENTAL AND SIMULATED RESULTS.

    Science.gov (United States)

    Pohorecki, Wladyslaw; Obryk, Barbara

    2017-09-29

    The results of nuclear heating measured by means of thermoluminescent dosemeters (TLD-LiF) in a Cu block irradiated by 14 MeV neutrons are presented. The integral Cu experiment relevant for verification of copper nuclear data at neutron energies characteristic for fusion facilities was performed in the ENEA FNG Laboratory at Frascati. Five types of TLDs were used: highly photon sensitive LiF:Mg,Cu,P (MCP-N), 7LiF:Mg,Cu,P (MCP-7) and standard, lower sensitivity LiF:Mg,Ti (MTS-N), 7LiF:Mg,Ti (MTS-7) and 6LiF:Mg,Ti (MTS-6). Calibration of the detectors was performed with gamma rays in terms of air-kerma (10 mGy of 137Cs air-kerma). Nuclear heating in the Cu block was also calculated with the use of MCNP transport code Nuclear heating in Cu and air in TLD's positions was calculated as well. The nuclear heating contribution from all simulated by MCNP6 code particles including protons, deuterons, alphas tritons and heavier ions produced by the neutron interactions were calculated. A trial of the direct comparison between experimental results and results of simulation was performed. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  19. The former tests realized to a personal neutron dosemeter based on solid nuclear tracks detector

    International Nuclear Information System (INIS)

    Camacho, M.E.; Tavera, L.; Balcazar, M.

    1997-01-01

    Due to the increase in the use of neutron radiation a personal neutron dosemeter based on solid nuclear tracks detector (DSTN) was designed and constructed. The personal dosemeter design consists of three arrangements. The first one consists of a plastic nuclear tracks detector (LR115 or CR39) in contact with a LiF pellet. The second one is the same that above but it placed among two cadmium pellets and, the third one is formed by the alone detector without converter neither neutron absorber. The three arrangements are placed inside a plastic porta detector hermetically closed to avoid the bottom produced by environmental radon whichever both detectors (LR115 and CR39) are sensitive. In this work the former tests realized to that dosemeter are presented. (Author)

  20. Abreu System - a dosimetric system to evaluate basic functioning parameters of roentgenography equipment

    International Nuclear Information System (INIS)

    Feital, J.C.

    1988-01-01

    This work shows a system to evaluate the half-value thickness of X-ray bundle. This system consists in a card with an aluminium filter, thermoluminescent dosemeter of lithium fluoride and radiographics films. )C.G.C.) [pt

  1. Investigations on low temperature thermoluminescence centres in quartz

    International Nuclear Information System (INIS)

    Bernhardt, H.

    1984-01-01

    The present paper will help to understand the often investigated process of thermoluminescence of quartz which is of high complexity. A lot of traps exist in quartz crystals which compete with each other with respect to the trapping of charge carriers during the X-ray treatment. That is why a variety of processes takes place after X-irradiation at liquid nitrogen temperature (LNT) of quartz which complicate the phenomenology of low temperature thermoluminescence. This competition in the trapping process leads to the so-called 'sensibilization' or 'desensibilization' effects of thermoluminescence, respectively, which are described in this paper for the first time. This effect means the dependence of the LNT thermoluminescence intensity on a pre-irradiation dose applied at room temperature (RT). The influence of this pre-irradiation is understood assuming the saturation of competitive traps. This favours an enhanced trapping of charge carriers at LNT-(shallow) traps instead of the preferential trapping on the deep traps in the case of X-ray treatment of the as-grown crystal at LNT. To get the afore mentioned model we take into account not only thermoluminescence but also coloration, ir- and vuv-absorption measurements. (author)

  2. Braided reinforced composite rods for the internal reinforcement of concrete

    Science.gov (United States)

    Gonilho Pereira, C.; Fangueiro, R.; Jalali, S.; Araujo, M.; Marques, P.

    2008-05-01

    This paper reports on the development of braided reinforced composite rods as a substitute for the steel reinforcement in concrete. The research work aims at understanding the mechanical behaviour of core-reinforced braided fabrics and braided reinforced composite rods, namely concerning the influence of the braiding angle, the type of core reinforcement fibre, and preloading and postloading conditions. The core-reinforced braided fabrics were made from polyester fibres for producing braided structures, and E-glass, carbon, HT polyethylene, and sisal fibres were used for the core reinforcement. The braided reinforced composite rods were obtained by impregnating the core-reinforced braided fabric with a vinyl ester resin. The preloading of the core-reinforced braided fabrics and the postloading of the braided reinforced composite rods were performed in three and two stages, respectively. The results of tensile tests carried out on different samples of core-reinforced braided fabrics are presented and discussed. The tensile and bending properties of the braided reinforced composite rods have been evaluated, and the results obtained are presented, discussed, and compared with those of conventional materials, such as steel.

  3. Thermoluminescence of meteorites and their orbits

    Science.gov (United States)

    Melcher, C. L.

    1981-01-01

    The thermoluminescence levels of 45 ordinary chondrites are measured in order to provide information on the orbital characteristics of the meteorites before impact. Glow curves of the photon emission response of powdered samples of the meteorites to temperatures up to 550 C in the natural state and following irradiation by a laboratory test dose of 110,000 rad were obtained as functions of terrestrial age and compared to those of samples of the Pribram, Lost City and Innisfree meteorites, for which accurate orbital data is available. The thermoluminescence levels in 40 out of 42 meteorites are found to be similar to those of the three control samples, indicating that the vast majority of ordinary chondrites that survive atmospheric entry have perihelia in the range 0.8-1 AU. Of the remaining two, Farmville is observed to exhibit an unusually large gradient in thermoluminescence levels with sample depth, which may be a result of a temperature gradient arising in a slowly rotating meteorite. Finally, the thermoluminescence measured in the Malakal meteorite is found to be two orders of magnitude lower than control samples, which is best explained by thermal draining by solar heating in an orbit with a perihelion distance of 0.5 to 0.6 AU.

  4. Dental tissue as a thermoluminescence dosimetry dosimeter

    International Nuclear Information System (INIS)

    Solaimani, F.; Zahmatkesh, M.H.; Akhlaghpoor, Sh.

    2003-01-01

    Background: Thermoluminescence dosimetry is one of the dosimetry procedures used widely as routine and personal dosimeters. In order to extend this kind of dosimeters, dental tissue has been examined and was found promising as a Thermoluminescence Dosimetry dosimeter. Materials and Methods: In this study, 70 health teeth were collected. The only criterion, wich was considered for selection of the teeth, was the healthiness of them regardless of age and gender of the donors. All collected samples were washed and cleaned and milled uniformly. The final powder had a uniform grain size between 100-300 micrometer. The sample was divided into four groups. Group A and B were used for measurement of density and investigation of variation of thermoluminescent characteristics with temperature respectively. Groups C and D were used for investigation of variation of thermoluminescent intensity with dose and fading of this intensity with time. In all cases the results obtained with dental tissue were compared to a standard LiF, thermoluminescence dosimetry dosimeter. Results: It was found that, average density of the dental tissue was 1.570 g/cm 3 , which is comparable to density of LiF, which is 1.612g/cm 3 . It was also concluded that the range of 0-300 d ig C , dental tissue has a simple curve with two specific peaks at 140 and 25 d ig C respectively. The experiment also showed that, the variation of relative intensity versus dose is linear in the range of 0.04-0.1 Gy. The fading rate of dental tissue is higher than LiF but still in the acceptable range (14% per month in compare to 5.2% per month). Conclusion: Dental tissue as a natural dosimeter is comparable with Thermoluminescence Dosimetry and can be used in accidental events with a good approximation

  5. Thermoluminescence - physics and applications

    International Nuclear Information System (INIS)

    Sunta, C.M.

    1992-01-01

    The interaction of ionizing radiations with matter leads to various effects, some of which leave memory in the target material. These memory effects can be seen as the after-effects of the irradiation. Thermoluminescence is one such phenomenon in the insulating solids. This paper describes the sequence of the physical events beginning with the incidence of an ionizing particle, leading to a trail of atomic and electronic displacements which stabilize in the form of so called defect centres. These defects - interstitials, colour centres, etc., store energy in their configuration. Most of the damage effects are reversible. The return of the displaced entities to their original site leads to the release of the stored energy. Thermoluminescence is the result of the re-adjustment (relaxation) of the displaced electrons. The stimulation for the relaxation process is provided by heating of the irradiated sample. The kinetics of the thermoluminescence process has been explained phenomenologically. The correspondence between the defect centres and the TL glow peaks has remained vague or ambiguous even in the most widely studied TL materials. Notwithstanding this deficiency, the phenomenon has found many practical applications. The paper deals mainly with the physical processes involved in the TL emission and alludes briefly to its involvement in applied areas.(author). 6 refs., 2 figs

  6. Effect of some parameters in the response of the Perspex 3042, Lot L amber dosemeters

    International Nuclear Information System (INIS)

    Prieto M, E.F.; Barrera G, G.

    2004-01-01

    The answer of the dosimetric systems is affected by several factors, for what should know as these factors they influence in each one of the different dosimetric systems and by this way to minimize its effect in the value of the absorbed dose and to obtain exact dose values. One of the dosimetric systems more used in the high dose dosimetry like routine dosemeter for the control of the irradiation process are the Perspex dosemeters, for their speed in the obtaining the information, their easy manipulation and the precision that they present. To this dosemeters group they belong the same as the Red and Clear the Amber, which are adequate for the measurement of the radiation dose in the range of high doses. The objective of the present work is to obtain the calibration curves of the dosemeters Amber Perspex 3042, Lot L under our work conditions, like they are the irradiation temperatures and of storage and to know the influence of the rate dose in the value of the absorbed dose for different measurement wavelengths, as well as, the relationship among the one post-irradiation time and the specific absorbance value induced in function of the absorbed dose. (Author)

  7. A portable versatile x-and γ-ray dosemeter with digital display for medical use

    International Nuclear Information System (INIS)

    Sankaran, A.; Gangadharan, P.

    1980-01-01

    A multipurpose dosemeter based on analogue-to-digital (A-D) conversion has been developed using solid state devices, micropower integrated circuits and displays. The A-D converter combines the features of a MOSFET electrometer, a voltage-to-frequency converter, the automatic Townsend balance, LED(LCD) display for exposure, analogue output and has a voltage-measuring accuracy of +-O.1% FS+-1 digit. The dosemeter which is designed to measure a wide range of X- and γ-ray exposure rates (1 mRh -1 to 1000 R min -1 ) and exposures (1 mR to 1000 R) with a variety of air-equivalent ionisation chambers, is compact, rugged and battery or AC mains powered. It is designed to provide trouble-free performance, easy operation and maintenance and to be independent of climatic conditions. Combining the features of a survey meter, dosemeter and a radioisotope calibrator, it can be fabricated using the facilities locally available in hospitals or universities. (author)

  8. Viability of the Fricke dosemeter doped with methylene blue

    International Nuclear Information System (INIS)

    Souza, V.L.B.; Santos, C.D.A.; Rodrigues, K.R.G.; Cunha, M.S.; Figueiredo, M.D.C.; Melo, R.T.

    2009-01-01

    This work aims to find the possible utilization of the Fricke dosemeter doped with methylene blue (FMB) for the dosimetry of photodynamic therapy. The FMB was irradiated wit X rays and light emitted diodes demonstrating positive answers to the stimulus, being probably to be used for dosimetric objectives

  9. Thermoluminescent dosimetry in pediatric patients subjected to TAC cranium multi cutting studies; Dosimetria termoluminiscente en pacientes pediatricos sometidos a estudios de TAC multicorte de craneo

    Energy Technology Data Exchange (ETDEWEB)

    Tabares V, M.J.; Azorin N, J. [UAM-Iztapalapa, 09340 Mexico D.F. (Mexico); Azorin V, J.C. [CICATA-IPN, 11500 Mexico D.F. (Mexico)]. e-mail: azorin@xanum.uam. mx

    2007-07-01

    Full text: In this work the doses to critical organs in pediatric patient subjected to studies of multi cutting skull tomography were determined. The study included patient smaller than 16. The used tomograph was a Siemens Somaton 16 plus multi cutting. The dosimetry one carries out with thermoluminescent dosemeters of LiF:Mg,Cu,P +Ptfe of national production. This dosemeters type was used due to it characteristics, such as low fading, equivalence with the tissue, easy handling and reading and independence of it answer with the radiation energy. The value of the absorbed doses in organs in the simple study results half of the obtained value in the contrasted study that which indicates that the CTDI{sub vol} is proportional to the dose to organ. In the case of the skull three-dimensional reconstruction tomography the CTDI{sub vol} was practically the same one that for simple tomography was practically; however, the doses to organs differed significantly. When comparing the TDI{sub vol} value of the three-dimensional reconstruction tomography with that of the contrasted tomography the obtained reason it was 1:2; however, the doses to organ didn't show the same relationship and the absorbed dose in the case of the thyroid gland was even bigger in the study of three-dimensional reconstruction. The value of the CTDI{sub vol} is a good indicator of the exposure factors selected for the tomographic exploration and it is useful for the determination of the effective dose, but it doesn't indicate the absorbed dose to organ. (Author)

  10. Exposure times and ratio of readings produced by irradiation of thermoluminescence dosemeters in PTB neutron standard fields between 0.1 MeV and 19 MeV

    International Nuclear Information System (INIS)

    Jahr, R.; Guldbakke, S.; Cosack, M.

    1979-06-01

    6 LiF and 7 LiF thermoluminescence dosimeters (TLD) irradiated in a standard field of monoenergetic neutrons (Esub(n) = 0.1 MeV to 19 MeV), respond as well to the photon as to the neutron component of this radiation field. The ratio of the two TLD readings as well as the duration of irradiation required to obtain a preset TLD reading are calculated as a function of the neutron energy. (orig.) [de

  11. A theoretical description of complex thermoluminescence curves: Pt. 1

    International Nuclear Information System (INIS)

    Kantorovich, L.N.; Fogel, G.M.; Gotlib, V.I.

    1988-01-01

    Thermoluminescence kinetics has been discussed within the framework of a band model assuming an extensive retrapping of charge carriers in comparison with their recombination. Assuming that a luminophor has one type of recombination centre and an arbitrary number of trapping centres we succeeded in analysing complex thermoluminescence curves with an arbitrary number of peaks. It has been shown for an elementary peak that its form almost coincides with the Randall-Wilkins curve, and its intensity is proportional to the square of the absorbed dose. The effect of both the processes of localised charge carriers accumulation (irradiation and isothermal decay) and their redistribution on trapping centres during the thermoluminescence process have been dealt with taking into account the form and intensity of the thermoluminescence peaks. (author)

  12. Neutron Dosimetry

    International Nuclear Information System (INIS)

    Vanhavere, F.

    2001-01-01

    The objective of SCK-CEN's R and D programme on neutron dosimetry is to improve the determination of neutron doses by studying neutron spectra, neutron dosemeters and shielding adaptations. In 2000, R and D focused on the contiued investigation of the bubble detectors type BD-PND and BDT, in particular their sensitivity and temperature dependence; the updating of SCK-CEN's criticality dosemeter, the investigation of the characteristics of new thermoluminescent materials and their use in neutron dosemetry; and the investigation of neutron shielding

  13. TELDE thermoluminescent dosimetry system

    International Nuclear Information System (INIS)

    Shvarts, K.K.; Nemiro, E.A.; Bichev, V.R.; Gotlib, V.I.; Grant, Z.A.; Grube, M.M.; Gubatova, D.J.

    1981-01-01

    The TELDE dosimetric set designed for measurement of x- and γ- rays doses with energy more than 10 keV is described. The set consists of thermoluminescent detectors from LiF (Ca, Mg, Ti). The detectors are polycrystalline pellets with diameter 3.5+-0.3 mm and thickness 2+-0.2 mm. In the thermoluminescence detectors both the peak and integral measurement methods are realized. Apart from this the TELDE set comprises the electron unit for the pre-irradiation thermal treatment of detectors, special bones for detectors storage and transportation, devices for their package in polyethelene film and containers to wear the detectors. The TELDE set allows to perform measurements on people or animals in radiobiological experiments as well as in water, solid or porous phantoms [ru

  14. Thermoluminescence spectra measured with a Michelson interferometer

    International Nuclear Information System (INIS)

    Haschberger, P.

    1991-01-01

    A Michelson interferometer was redesigned to prove its capabilities in the measurement of short-lived, low-intensity thermoluminescence spectra. Interferograms are collected during heating up the thermoluminescent probe in a heater plate. A personal computer controls the data acquisition and processes the Fourier transform. As the results show, even a comparatively simple and limited setup leads to relevant and reproducible spectra. (author)

  15. Study of a selection of 10 historical types of dosemeter: variation of the response to Hp(10) with photon energy and geometry of exposure.

    Science.gov (United States)

    Thierry-Chef, I; Pernicka, F; Marshall, M; Cardis, E; Andreo, P

    2002-01-01

    An international collaborative study of cancer risk among workers in the nuclear industry is tinder way to estimate direetly the cancer risk following protracted low-dose exposure to ionising radiation. An essential aspect of this study is the characterisation and quantification of errors in available dose estimates. One major source of errors is dosemeter response in workplace exposure conditions. Little information is available on energy and geometry response for most of the 124 different dosemeters used historically in participating facilities. Experiments were therefore set up to assess this. using 10 dosemeter types representative of those used over time. Results show that the largest errors were associated with the response of early dosemeters to low-energy photon radiation. Good response was found with modern dosemeters. even at low energy. These results are being used to estimate errors in the response for each dosemeter type, used in the participating facilities, so that these can be taken into account in the estimates of cancer risk.

  16. Thermoluminescence in alkali halides irradiated at 80K

    International Nuclear Information System (INIS)

    Jimenez de Castro, M.

    1978-01-01

    The thermoluminescence, the thermally stimulated currents and the thermal stability of the F centres induced in pure NaCl and KC1 crystals by X irradiation at 80K have been studied in detail, in the range between 80 and 300K. The thermoluminescent processes induced by illumination at 80K with F light in samples previously irradiated at room temperature has also been studied. It has been clearly observed the existence of thermoluminescent processes due to electrons and holes thermally released from traps, in which the F centres are not involved. The existence of hole-F centre recombination has not been observed. There are several thermoluminescent processes in both materials which are scribed to the recombination of F centres with mobile interstitial halogen atoms thermally released from traps, which are likely monovalent impurities in this temperature interval. The light emitting stage in these processes is originated by the formation of self trapped excitons. (Author) 66 refs

  17. Equipment for observing thermoluminescence

    International Nuclear Information System (INIS)

    Sette Camara, Alexandre; Cogneau, Michel; Ristori, Christian; Vargas, Jose Israel

    1971-06-01

    A detailed description is given of an apparatus for investigating thermoluminescence properties. Chopping of the emitted light and subsequent amplification of the electric signal derived led to high stability and sensitivity [fr

  18. Self-sensing CF-GFRP rods as mechanical reinforcement and sensors of concrete beams

    Science.gov (United States)

    Nanni, F.; Auricchio, F.; Sarchi, F.; Forte, G.; Gusmano, G.

    2006-02-01

    In this paper testing carried out on concrete beams reinforced with self-sensing composite rods is presented. Such concrete beams, whose peculiarity is to be reinforced by self-sensing materials able to generate an alarm signal when fixed loads are reached, were designed, manufactured and tested. The reinforcing rods were manufactured by pultrusion and consisted of self-sensing hybrid composites containing both glass and carbon fibres in an epoxy resin. The experimentation was carried out by performing simultaneously mechanical tests on the reinforced beams and electrical measurements on the composite rods. The results showed that the developed system reached the target proposed, giving an alarm signal.

  19. Study on routines and procedures of calibration of clinical dosemeter on gamma radiation beams at the IPEN-CNEN/SP, Brazil

    International Nuclear Information System (INIS)

    Damatto, Willian B.; Potiens, Maria da Penha A.; Santos, Gelson P.; Vivolo, Vitor

    2009-01-01

    The calibration consists in the comparison of the equipment to be calibrated with other equipment already calibrated by another standard, which means the equipment that will accomplish shall be considered secondary or tertiary, depending on the calibration that will be accomplished. As the clinical dosemeters are used for quality control services in the hospitals, these dosemeters need a reliable measurement precision. Therefore, this work intends to demonstrate the importance of the clinical dosemeter and the calibration, requiring special care face to all clinical set

  20. Individual Dosimetry for High Energy Radiation Fields

    International Nuclear Information System (INIS)

    Spurny, F.

    1999-01-01

    The exposure of individuals on board aircraft increased interest in individual dosimetry in high energy radiation fields. These fields, both in the case of cosmic rays as primary radiation and at high energy particle accelerators are complex, with a large diversity of particle types, their energies, and linear energy transfer (LET). Several already existing individual dosemeters have been tested in such fields. For the component with high LET (mostly neutrons) etched track detectors were tested with and without fissile radiators, nuclear emulsions, bubble detectors for both types available and an albedo dosemeter. Individual dosimetry for the low LET component has been performed with thermoluminescent detectors (TLDs), photographic film dosemeters and two types of electronic individual dosemeters. It was found that individual dosimetry for the low LET component was satisfactory with the dosemeters tested. As far as the high LET component is concerned, there are problems with both the sensitivity and the energy response. (author)

  1. Hysteresis, reentrance, and glassy dynamics in systems of self-propelled rods.

    Science.gov (United States)

    Kuan, Hui-Shun; Blackwell, Robert; Hough, Loren E; Glaser, Matthew A; Betterton, M D

    2015-01-01

    Nonequilibrium active matter made up of self-driven particles with short-range repulsive interactions is a useful minimal system to study active matter as the system exhibits collective motion and nonequilibrium order-disorder transitions. We studied high-aspect-ratio self-propelled rods over a wide range of packing fractions and driving to determine the nonequilibrium state diagram and dynamic properties. Flocking and nematic-laning states occupy much of the parameter space. In the flocking state, the average internal pressure is high and structural and mechanical relaxation times are long, suggesting that rods in flocks are in a translating glassy state despite overall flock motion. In contrast, the nematic-laning state shows fluidlike behavior. The flocking state occupies regions of the state diagram at both low and high packing fraction separated by nematic-laning at low driving and a history-dependent region at higher driving; the nematic-laning state transitions to the flocking state for both compression and expansion. We propose that the laning-flocking transitions are a type of glass transition that, in contrast to other glass-forming systems, can show fluidization as density increases. The fluid internal dynamics and ballistic transport of the nematic-laning state may promote collective dynamics of rod-shaped micro-organisms.

  2. Performance of Panasonic ZP-1460 Electronic Personal Dosemeter under Exposure Conditions Likely to be Found at Fukushima Daiichi Nuclear Power Plant

    Science.gov (United States)

    Tsujimura, Norio; Yoshida, Tadayoshi; Hoshi, Katsuya; Momose, Takumaro

    A study on the performance of the Panasonic ZP-1460 electronic personal dosemeter, the model used in the aftermath of the Fukushima Daiichi nuclear power plant accident in March 2011, was conducted under actual exposure situations likely encountered in the plant. The tests pertained to (1) the dose rate response over dose rates >100 mSv/h and (2) the angular response on an anthropomorphic phantom exposed to the rotational and isotropic irradiation geometries. The test results confirmed that the dosemeter provides Hp(10) as a reasonably close estimate of the effective dose for any exposure geometries. The dosemeter response data evaluated in this study can be utilized for converting dosemeter readings to the absorbed dose to any organs and tissues for epidemiologic purposes.

  3. Tensile Characterization of FRP Rods for Reinforced Concrete Structures

    Science.gov (United States)

    Micelli, F.; Nanni, A.

    2003-07-01

    The application of FRP rods as an internal or external reinforcement in new or damaged concrete structures is based on the development of design equations that take into account the mechanical properties of FRP material systems.The measurement of mechanical characteristics of FRP requires a special anchoring and protocol, since it is well known that these characteristics depend on the direction and content of fibers. In this study, an effective tensile test method is described for the mechanical characterization of FRP rods. Twelve types of glass and carbon FRP specimens with different sizes and surface characteristics were tested to validate the procedure proposed. In all, 79 tensile tests were performed, and the results obtained are discussed in this paper. Recommendations are given for specimen preparation and test setup in order to facilitate the further investigation and standardization of the FRP rods used in civil engineering.

  4. Thermoluminescent phosphor

    Science.gov (United States)

    Lasky, Jerome B.; Moran, Paul R.

    1978-01-01

    A thermoluminescent phosphor comprising LiF doped with boron and magnesium is produced by diffusion of boron into a conventional LiF phosphor doped with magnesium. Where the boron dopant is made to penetrate only the outer layer of the phosphor, it can be used to detect shallowly penetrating radiation such as tritium beta rays in the presence of a background of more penetrating radiation.

  5. Effect of co-doping of sodium on the thermoluminescence dosimetry properties of copper-doped zinc lithium borate glass system

    International Nuclear Information System (INIS)

    Saidu, A.; Wagiran, H.; Saeed, M.A.; Alajerami, Y.S.M.; Kadir, A.B.A.

    2016-01-01

    The effect of sodium as a co-dopant on the thermoluminescence (TL) properties of copper-doped zinc lithium borate (ZLB: Cu) subjected to Co-60 gamma radiation is reported in this study. TL intensity is enhanced with the introduction of sodium in ZLB: Cu. The obtained glow curve is simple with a single peak. The annealing procedure and the best heating rate for the proposed thermoluminescent dosimeter (TLD) are established, and the phosphor is reusable. The TL response within the dose range of 0.5–1000 Gy is investigated. The results show that the thermal fading behaviour is improved significantly. - Highlights: • Dosimetry properties of an improved TL dosimeter. • The dosimeter is made of lithium borate, modified with ZnO, doped with CuO and co-doped with Na 2 O. • With addition of Na to Cu in the ZLB host, TL yield and sensitivity has significantly enhanced. • The fading behaviour has also been minimized significantly. • The new material is also characterized with the linear dose response, and good reproducibility behaviour.

  6. Calibration of a special neutron dosemeter based on solid-state track detectors and fission radiators in various neutron fields

    International Nuclear Information System (INIS)

    Doerschel, B.; Krusche, M.; Schuricht, V.

    1980-01-01

    The calibration of a personnel neutron dosemeter in different neutron fields is described. The badge-like dosemeter contains 5 detectors: polycarbonate foil (10 μm, Makrofol KG), 232 Th, natural uranium, natural uranium with boron, and natural uranium with cadmium. Detector sensitivity and calibration factors have been calculated and measured in radiation fields of 252 Cf fission neutrons, WWR-S reactor neutrons with and without Cd and Fe shielding, 3-MeV (d,t) generator neutrons, and 238 PuBe neutrons. Measurement range and achievable accuracy are discussed from the point of view of applying the dosemeter in routine and emergency uses

  7. Graphite mixed magnesium borate TL dosemeters for beta ray dosimetry

    DEFF Research Database (Denmark)

    Prokic, M; Christensen, Poul

    1984-01-01

    Sintered MgB4O7:Dy dosemeters with graphite contents from 1 to 10% were investigated for application for personnel dosimetry. Data are given on dose response, dose threshold, reproducibility, beta energy response and fading. Furthermore, results from practical field experiments are presented...

  8. Radiation accident dosimetry: TL properties of mobile phone screen glass

    International Nuclear Information System (INIS)

    Bassinet, C.; Pirault, N.; Baumann, M.; Clairand, I.

    2014-01-01

    Mobile phones are carried by a large part of the population and previous studies have shown that they may be able to function as individual fortuitous dosimeters in case of radiological accident. This study deals with thermoluminescence (TL) properties of mobile phone screen glass. The presence of a significant background signal which partially overlaps with the radiation-induced signal is a serious issue for dose reconstruction. A mechanical method to reduce this signal using a diamond grinding bit is presented. An over-response at low energy (∼50 keV) is observed for two investigated glasses. The results of a dose recovery test using a single-aliquot regenerative-dose (SAR) procedure are discussed. - Highlights: • Mobile phone screen glass is a promising material for retrospective dosimetry. • The TL non-radiation induced background signal can be significantly reduced by a mechanical method. • A dose recovery test using an SAR procedure was successfully carried out for the investigated glass

  9. Annual radiation dose in thermoluminescence dating

    International Nuclear Information System (INIS)

    Li Huhou

    1988-01-01

    The annual radiation dose in thermoluminescence dating has been discussed. The autor gives an entirely new concept of the enviromental radiation in the thermoluminescence dating. Methods of annual dose detemination used by author are dating. Methods of annual dose determination used by author are summed up, and the results of different methods are compared. The emanium escapiug of three radioactive decay serieses in nature has been considered, and several determination methods are described. The contribution of cosmic rays for the annual radiation dose has been mentioned

  10. Annual radiation dose in thermoluminescence dating

    Energy Technology Data Exchange (ETDEWEB)

    Huhou, Li [Chinese Academy of Social Sciences, Beijing, BJ (China). Inst. of Archaeology

    1988-11-01

    The annual radiation dose in thermoluminescence dating has been discussed. The autor gives an entirely new concept of the enviromental radiation in the thermoluminescence dating. Methods of annual dose detemination used by author are dating. Methods of annual dose determination used by author are summed up, and the results of different methods are compared. The emanium escapiug of three radioactive decay serieses in nature has been considered, and several determination methods are described. The contribution of cosmic rays for the annual radiation dose has been mentioned.

  11. Development of Thermoluminescence Dosimeter CaSO4:Dy as Personal and Environmental Dosimeters

    International Nuclear Information System (INIS)

    Hasnel Sofyan

    2009-01-01

    Development of personal and environmental dosimeters using material phosphors of CaSO 4 :Dy powder in form capillary glass and disc teflon thermoluminescence (TL) dosimeter have been done. TL dosimeter CaSO 4 :Dy powder used can record dose response less than 0.01 mGy. Fading of TL dosimeter capillary glass after 29 days is 25%. In 1 batch, making of CaSO 4 :Dy powder can obtain 2 groups of dosimeter capillaries with coefficient variance smaller than 10%. This discrepancy caused difference in powder making and reading of the TL dosimeter. TL dosimeter CaSO 4 :Dy teflon disc with dia. 5 mm and 0.8 mm thickness is homogeneous mixture between phosphor powder with dia. 80 to 150 mesh and teflon powder dia. 20 μm. The composition of CaSO 4 :Dy and teflon in TL dosimeter influence sensitivity of the dosimeter. It’ concluded that in order to obtain optimal sensitivity of TL dosimeter, the composition of CaSO 4 :Dy and teflon is 3 and 1 with pressured of disc in 700 MPa. (author).

  12. Evaluation of fading factor and self-dose for glass dosimeter and thermoluminescence dosimeter

    International Nuclear Information System (INIS)

    Yamasaki, T.; Yamanishi, H.; Miyake, H.; Komura, K.

    2000-01-01

    The glass dosimeter (GD) and thermoluminescence dosimeter (TLD) are both passive radiation detectors. They are often used for measuring environmental radiation. In order to measure low dose rate preciously, it is important to evaluate decreased dose due to fading and self-dose during the exposure period. We evaluate the fading factor and self-dose of thee passive detectors, GD and TLD. We select Ogoya tunnel for the experiment. The tunnel is suitable field for measuring faded dose and self-dose because it is low cosmic radiation. At the center of the tunnel, the intensity of cosmic ray is reduced to about 1/177 than the outside of the funnel. We prepared two sets of dosimeters. One set consists of five GDs, five TLDs and some pre-irradiated GDs and TLDs that are exposed to standard radiation of 4 mGy by Cs-137. These dosimeters are put in the 10 cm thick lead box in order to shield the terrestrial gamma ray. One set is located at the center of the tunnel and the other is the outside of the funnel. The dosimeters were exposed for ten months, from May 1998 to March 1999. After the exposure, the readers of dosimeters are carried into the funnel to read out the signals promptly as soon as taking out the dosimeters. As a result of the measurement, four kinds of data are taken for GD and TLD respectively. Assumed that the self-dose and cosmic ray are constant during exposure, the four independent unknown quantities, a self-dose a dose due to cosmic ray and a fading coefficient at the center of the tunnel and at the outside, are considered. Therefore four simultaneous equations should be obtained. From these examinations, the faded dose of GD is less than 1%, but that of TLD is about 16% during ten months. The coefficient for compensation of fading of GD and TLD is given as the half of the each value. At the outside of the tunnel, the measured dose rate of cosmic ray that can pass through the 10 cm lead is evaluated to be about 16 nGy/h by both detectors. The self

  13. Neutron Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Vanhavere, F

    2001-04-01

    The objective of SCK-CEN's R and D programme on neutron dosimetry is to improve the determination of neutron doses by studying neutron spectra, neutron dosemeters and shielding adaptations. In 2000, R and D focused on the contiued investigation of the bubble detectors type BD-PND and BDT, in particular their sensitivity and temperature dependence; the updating of SCK-CEN's criticality dosemeter, the investigation of the characteristics of new thermoluminescent materials and their use in neutron dosemetry; and the investigation of neutron shielding.

  14. Determination of isodose curves in Radiotherapy using an Alanine/ESR dosemeter; Determinacion de curvas de isodosis en Radioterapia usando un dosimetro de Alanina/ESR

    Energy Technology Data Exchange (ETDEWEB)

    Chen, F.; Baffa, O.; Graeff, C.F.O. [Departamento de Fisica e Matematica. Universidade de Sao Paulo FFCLRP. 14040-901 Ribeirao Preto-SP (Brazil)

    1998-12-31

    It was studied the possible use of an Alanine/ESR dosemeter in the isodose curves mapping in normal treatments of Radiotherapy. It was manufactured a lot of 150 dosemeters with base in a mixture of D-L Alanine dust (80 %) and paraffin (20 %). Each dosemeter has 4.7 mm diameter and 12 mm length. A group of 100 dosemeters of the lot were arranged inside 50 holes of the slice 25 of the phantom Rando Man. The phantom irradiation was realized in two opposed projections (AP and PA) in Co-60 equipment. A group of 15 dosemeters was take of the same lot for obtaining the calibration curve in a 1-20 Gy range. After irradiation the signal of each dosemeter was measured in an ESR spectrometer operating in the X-band ({approx} 9.5 GHz) and the wideness of Alanine ESR spectra central line was correlated with the radiation dose. The wideness dose calibration curve resulted linear with a correlation coefficient 0.9996. The isodose curves obtained show a profile enough similar at comparing with the theoretical curves. (Author)

  15. Systematic errors in the readings of track etch neutron dosemeters caused by the energy dependence of response

    International Nuclear Information System (INIS)

    Tanner, R.J.; Thomas, D.J.; Bartlett, D.T.; Horwood, N.

    1999-01-01

    A study has been performed to assess the extent to which variations in the energy dependence of response of neutron personal dosemeters can cause systematic errors in readings obtained in workplace fields. This involved a detailed determination of the response functions of personal dosemeters used in the UK. These response functions were folded with workplace spectra to ascertain the under- or over-response in workplace fields

  16. Systematic errors in the readings of track etch neutron dosemeters caused by the energy dependence of response

    CERN Document Server

    Tanner, R J; Bartlett, D T; Horwood, N

    1999-01-01

    A study has been performed to assess the extent to which variations in the energy dependence of response of neutron personal dosemeters can cause systematic errors in readings obtained in workplace fields. This involved a detailed determination of the response functions of personal dosemeters used in the UK. These response functions were folded with workplace spectra to ascertain the under- or over-response in workplace fields.

  17. Use of a robotic manipulator in the simulation of the automation of a calibration process of dosemeters

    International Nuclear Information System (INIS)

    Benitez R, J.S.; Najera H, M.C.

    2002-01-01

    The development of a system based in a manipulative robot which simulates the operative sequence in a calibration process of dosemeters is presented. In this process it is performed the monitoring of the dosemeter positions and the calibrator by mean of an arm of articulated robot which develops the movement sequences and the taking a decision based on the information coming from the external sensors. (Author)

  18. Application of thermoluminescence measurements to detect irradiated strawberries

    International Nuclear Information System (INIS)

    Heide, L.; Guggenberger, R.; Bogl, K.W.

    1990-01-01

    The thermoluminescence intensity of unirradiated and irradiated strawberries was studied with regard to dose response, storage time, and different varieties. An identification method could finally be developed. Further investigations were carried out to determine the origin of the thermoluminescence effect, which was found to be attributable to mineral grains adhering to the sample surface

  19. A consolidation process for spent burnable poison rod assemblies

    International Nuclear Information System (INIS)

    Yamamoto, Y.; Harada, M.; Komatsu, Y.

    1985-01-01

    A new consolidation system for the spent burnable poison assembly utilizing a sequence control robot operated under water was proposed. A credible accident in the system was analyzed mainly from the viewpoint of tritium release, based on the diffusion analysis of tritium in borosilicate glass. It was found that the amount of tritium released would be small even after the rupture of burnable poison rods. An experiment on a new consolidation system was performed using spent burnable poison assemblies. The volume of burnable poison assemblies was reduced safely and securely by a factor of 7 to 14 for burnable poison rods and by 22 for hold-down portions. It was proved that the consolidation system is collectively feasible

  20. Manufacture of thermoluminescent dosimeter of LiF(Mg) by sintherization technique

    International Nuclear Information System (INIS)

    Borges, V.; Elbern, A.W.

    1984-01-01

    A technique to produce solid state dosemeters, that can be used in personal dosimetry, using LiF powder is described. Many differents baths of cold compacted and sinterized dosimeters were investigated as a function of the dopant concentration and the sinterization time. Some termoluminescent characteristics were measured, and compared to those of the extruded TLD-100 dosemeters, which are used as an international reference. The experimental methodology and the obtained results, with their respective analyses, are the subject of this paper. (M.A.C.) [pt

  1. Influence of the PMMA and the ISO slab phantom for calibrating personal dosemeters

    International Nuclear Information System (INIS)

    Ginjaume, M.; Ortega, X.; Barbosa, A.

    2000-01-01

    Wide agreement has been achieved among the main Bodies and Organisations involved in standardisation of radiation protection, related to the operational quantities for personal dosimetry. Since their definition in 1985, several reports have been published to clarify and define the experimental set up to be used for the calibration of dosemeters in terms of the above-mentioned quantities. Among these lines, ICRU 47 has listed five different phantoms that are used for calibration and whose results were accurate, within accepted uncertainties. However, to achieve uniformity in calibration procedures, the 30 cm x 30 cm x 15 cm PMMA slab phantom was recommended. The secondary calibration Laboratory from the Institut de Techniques Energetiques at the Technical University of Catalonia (UPC) agreed with the Spanish Nacional Laboratory to adopt the recommended PMMA phantom but to improve the accuracy of the calibration procedure, by introducing a correction factor for backscatter differences in a PMMA and an ICRU slab phantom. Such corrections were of the order of 8% for the low-energy X-ray qualities. Recently, ISO in ISO 4037-3, has proposed the ISO water slab phantom which consists of a 30 cm x 30 cm x 15 cm water phantom with PMMA walls (front wall 2.5 mm thick). This new phantom could be substituted by the above mentioned PMMA phantom for radiation qualities with mean energy equal or above that of 137 Cs. The aim of this work is to compare the influence of both phantoms when calibrating personal dosemeters with photons. A set of four TL personal dosemeters that are used in the UPC personal dosimetry Service and an electronic personal dosemeter (SIEMENS EPD-2) were calibrated in terms of H p (10) and H p (0.07) using the two proposed phantoms. Calibration factors for ISO X-ray narrow spectra, 137 Cs and 60 Co were experimentally obtained for each phantom and compared. In the TL measurements, differences were found to be within TL statistical uncertainty, provided that a

  2. Thermoluminescence in films of HfO2:Dy+3

    International Nuclear Information System (INIS)

    Ceron, P.; Rivera, T.; Guzman, J.; Montes, E.; Pelaez, A.; Rojas, B.; Guzman, D.; Azorin, J.; Paredes, L.

    2014-08-01

    In this work the thermoluminescence (TL) response of films of hafnium oxide polluted with dysprosium (HfO 2 :Dy +3 ) that were irradiated in the near UV (200 nm - 400 nm). The films were deposited by means of the ultrasonics spray pyrolysis technique on a glass substrate, using different deposit temperatures (300 grades C - 600 grades C). The best TL emission corresponded to the prepared film to 450 grades C that was exposed to a spectral irradiation of 80 μJ/(cm 2 -s) with a wave longitude of 240 nm. The TL response in function of the spectral irradiation was lineal in the studied interval (24 to 288 mJ/cm 2 ), several kinetic parameters were also calculated of the shine curve as depth of the trap (E), frequency factor (s) and order to the kinetics (b). The obtained results show that the films of HfO 2 :Dy +3 could be used as radiation monitor in the region of the near UV. (Author)

  3. Use of bubbles dosemeters to measure the neutrons leakage radiations around a medical accelerator

    International Nuclear Information System (INIS)

    Bourgois, L.

    1999-01-01

    In this study, the neutron component has been determined on a medical accelerator (Saturne 43F type accelerator) with the help of bubbles detectors marketed by the Bubble Technology Industry society. BD-PND type dosemeters with sensitivities from 0.1 to 0.23 bubbles/micro sievert have been used. These dosemeters have important advantages: they are small enough and then, homogeneously irradiated, they are insensitive to intense electromagnetic fields and to photons and electrons, they can used several times by compression, the reading of the device does not need expensive means and it is possible to multiply the measure points. (N.C.)

  4. Thermoluminescence, luminescence optically stimulated and creation of defects in alkaline halogenides contaminated with Europium; Termoluminiscencia, luminiscencia opticamente estimulada y creacion de defectos en halogenuros alcalinos contaminados con Europio

    Energy Technology Data Exchange (ETDEWEB)

    Barboza F, M. [Centro de Investigacion en Fisica, Universidad de Sonora, Apartado Postal 5-088, 83190 Hermosillo, Sonora (Mexico)

    1999-07-01

    The alkaline halogenides have been subject matter of investigations related with the search of sensor materials for X-ray bidimensional images or optical memories. The understanding of the damage formation processes generated by ionizing and non-ionizing radiations is important for the correct design of devices that working as detectors and dosemeters of both type of radiations. In this work we present the investigation results related with the defects produced by the ionizing radiation type X and ultraviolet light in the range of 200-360 nm in crystals of KCl: Eu{sup 2+} and KBr:Eu{sup 2+}. It is examined the thermoluminescence and luminescence spectra with the purpose of identifying the exciton processes, owing to the excitation of the halogenide ions in which the primary defects correspond to the F and H centers. It has been found that the 400-600 nm emission is associated with the luminescence type that in his turn can be associated with autotrapped excitons perturbed by the impurity. On the other hand, it is examined the emission spectra of the luminescence optically stimulated in crystals of KBr: Eu{sup 2+} and KCl: Eu{sup 2+} finding too that such materials would be used as optical memories susceptible of to store information, and through of photostimulation to read this. It was determined that the F centers participate in the luminescence optically stimulated in these crystals, as well as too in the recombination processes responsible by the thermoluminescent emission. (Author)

  5. Dose measurements in space by the Hungarian Pille TLD system

    International Nuclear Information System (INIS)

    Apathy, I.; Deme, S.; Feher, I.; Akatov, Y.A.; Reitz, G.; Arkhanguelski, V.V.

    2002-01-01

    Exposure of crew, equipment, and experiments to the ambient space radiation environment in low Earth orbit poses one of the most significant problems to long-term space habitation. Accurate dose measurement has become increasingly important during the assembly (extravehicular activity (EVA)) and operation of space stations such as on Space Station Mir. Passive integrating detector systems such as thermoluminescent dosemeters (TLDs) are commonly used for dosimetry mapping and personal dosimetry on space vehicles. The well-known advantages of passive detector systems are their independence of power supply, small dimensions, high sensitivity, good stability, wide measuring range, resistance to environmental effects, and relatively low cost. Nevertheless, they have the general disadvantage that for evaluation purposes they need a laboratory or large--in mass and power consumption--terrestrial equipment, and consequently they cannot provide time-resolved dose data during long-term space flights. KFKI Atomic Energy Research Institute (KFKI AEKI) has developed and manufactured a series of thermoluminescent dosemeter systems for measuring cosmic radiation doses in the 10 μGy to 10 Gy range, consisting of a set of bulb dosemeters and a compact, self-contained, TLD reader suitable for on-board evaluation of the dosemeters. By means of such a system, highly accurate measurements were carried out on board the Salyut-6, -7 and Mir Space Stations as well as on the Space Shuttle. A detailed description of the system is given and the comprehensive results of these measurements are summarised

  6. Improvements in the Tl dosimetry for Radiotherapy; Mejoras en la dosimetria Tl para Radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Bustos, S.; Velez, G.; Rubio, M. [CEPROCOR. Arenales 230 Juniors Cordoba 5000 (Argentina)

    1998-12-31

    The thermoluminescent dosimetry (TLD) in vivo has been demonstrated to be one of the most reliable for the control of radiotherapeutic treatments, but the delay in the response is the main disadvantage in its applicability. In this work are presented important improvements and it is demonstrated that maintaining the accuracy and reliability of the technique, it is possible to accelerate the response times at a few hours. To realize this work is utilized a lecturer Harshaw 4000, dosemeters LiF-TLD-100 chips (3.1 x 3.1 x 0.89 mm{sup 3}) and rods (1 x 1 x 6 mm{sup 3}). With the implementation of a glow curve analysis program developed in CIEMAT, its obtained a Tl peaks separation in such manner rapid and accurate, by that the thermal treatment of dosemeters may be reduced at one unique annealing pre-irradiation for 1 hr at 400 Centigrade. It is realized a periodical and individual calibration of the TLD and a study of the factors which influencing the ratio Tl signal-dose as linearity, correction by energy, directional response and pride of Tl signal. the results of this study are introducing in a calculation list specially designed and which allows to obtain absorbed dose by TLD starting of the dates (dosimetric peaks area) which appear of the glow curves analysis. The dose is obtained with an accuracy less than 5 %. The dosemeters already irradiated (in vivo) are analysed and informed in only four hours, allowing a greater control of the treatment and a correction of the possible errors for the next session, still in bi fractional treatments. The method implemented results thus accurate, rapid and reliable. (Author)

  7. Improvements in the Tl dosimetry for Radiotherapy

    International Nuclear Information System (INIS)

    Bustos, S.; Velez, G.; Rubio, M.

    1998-01-01

    The thermoluminescent dosimetry (TLD) in vivo has been demonstrated to be one of the most reliable for the control of radiotherapeutic treatments, but the delay in the response is the main disadvantage in its applicability. In this work are presented important improvements and it is demonstrated that maintaining the accuracy and reliability of the technique, it is possible to accelerate the response times at a few hours. To realize this work is utilized a lecturer Harshaw 4000, dosemeters LiF-TLD-100 chips (3.1 x 3.1 x 0.89 mm 3 ) and rods (1 x 1 x 6 mm 3 ). With the implementation of a glow curve analysis program developed in CIEMAT, its obtained a Tl peaks separation in such manner rapid and accurate, by that the thermal treatment of dosemeters may be reduced at one unique annealing pre-irradiation for 1 hr at 400 Centigrade. It is realized a periodical and individual calibration of the TLD and a study of the factors which influencing the ratio Tl signal-dose as linearity, correction by energy, directional response and pride of Tl signal. the results of this study are introducing in a calculation list specially designed and which allows to obtain absorbed dose by TLD starting of the dates (dosimetric peaks area) which appear of the glow curves analysis. The dose is obtained with an accuracy less than 5 %. The dosemeters already irradiated (in vivo) are analysed and informed in only four hours, allowing a greater control of the treatment and a correction of the possible errors for the next session, still in bi fractional treatments. The method implemented results thus accurate, rapid and reliable. (Author)

  8. Thermoluminescent behavior of diamond thin films exposed to ultraviolet radiation

    International Nuclear Information System (INIS)

    Barboza F, M.; Gastelum, S.; Melendrez, R.; Chernov, V.; Bernal, R.; Cruz V, C.; Brown, F.

    2002-01-01

    In this work the thermoluminescent properties of diamond thin films are discussed which are grown up through the chemical vapor method exposed to ultraviolet radiation of 200-280 nm. The films with thickness 3, 6, 9, 12, 180 and 500 microns were grown up using a precursor gas formed of H 2 -CH 4 -CO excited through microwave energy or hot filament.The structure and morphology of the films were examined through scanning electron microscopy, indicating the formation of different diamond polycrystal structures which depend on the type of heating of the precursor gas used as well as the film dimensions. In general, the brilliance curve depends on the sample and the wavelength of the irradiation ultraviolet light, however it presents clearly thermoluminescence bands in 148, 160, 272, 304, 320 and 324 C degrees. The maximum of the thermoluminescence efficiency is obtained for the case of sample exposure with light of 214 nm. The sample of 500 microns is what exhibits greater thermoluminescent efficiency of those studied samples. The thermoluminescent behavior in function of radiation dose presents regions of linearity and supra linearity for higher and small doses respectively. The disappearance of the thermoluminescent signal depends on the characteristics of the film and it can reach until a 30 % of loss before to reach the stability. (Author)

  9. Characterization of workplaces in interventional radiology using active dosemeters ALARA OD

    International Nuclear Information System (INIS)

    Prlic, I.; Milkovic-Kraus, S.; Mestrovic, T.; Suric-Mihic, M.; Vrtar, M.

    2005-01-01

    Full text: Because of progressive development and extended use of interventional radiology procedures it is highly recommended that all individuals involved in the process should be aware of the potential for both stochastic and deterministic effects due to occupational exposure. Interventional radiology procedures are essentially therapeutic and are performed by various medical specialists, not only by properly educated radiologists. As the procedures are performed in such a manner that certain number of medical staff are always needed near the patient, near the x- ray unit, ensuring 'safe' working environment in such radiation x-ray field geometry is a new challenge to regular radiation protection. In this work we are not primary concerned in relatively high doses delivered to patients undergoing interventional procedures. The patient is rather regarded as a secondary radiation source which emits scattered x-rays. The working staff, moving in mixed, primary and scattered x-ray field, is expected to be exposed to higher occupational doses due to combination of extended fluoroscopy times, elevated fluoroscopy currents and larger amount of radiographic images required. The protection of both patients and staff is to be upgraded. As passive dosemetes will give us clear knowledge only about the monthly integrated occupational dose, a reasonable doubt exists about the frequency and duration, of receiving the dose. Measuring dose rate is not a part of regular passive dosemeters monitoring systems. This is why we have developed active - electronic dosemeter device which provides us with additional dosimetry data about the frequency and duration of professional exposure burden. Digital dosemeter ALARA OD will record and integrate any occupational dose including the normal radiation background in the working area. It will record the time and duration of any fluoroscopic exposure done. This gives us the data about the dose rate of occupational radiation and frequency of

  10. The radioprotection inspection manual for odontologic X-rays equipment

    International Nuclear Information System (INIS)

    Almeida, C.D. de.

    1988-01-01

    The inspection manual for odontologic X-ray, using films and thermoluminescent dosemeters are showed, decreasing the time of exposition by some recommendation. Some calculations are also cited, for decrease the exposure time, for filtrating the equipment and for a radiometric survey. (C.G.C.) [pt

  11. Control rod

    International Nuclear Information System (INIS)

    Kawakami, Kazuo; Shimoshige, Takanori; Nishimura, Akira

    1979-01-01

    Purpose: A control rod has been developed, which provided a plurality of through-holes in the vicinity of the sheath fitting position, in order to flatten burn-up, of fuel rods in positions confronting a control rod. Thereby to facilitate the manufacture of the control rods and prevent fuel rod failures. Constitution: A plurality of through-holes are formed in the vicinity of the sheath fitting position of a central support rod to which a sheath for the control rod is fitted. These through-holes are arranged in the axial direction of the central support rod. Accordingly, burn-up of fuel rods confronting the control rods can be reduced by through-holes and fuel rod failures can be prevented. (Yoshino, Y.)

  12. Eye lens monitoring for interventional radiology personnel: dosemeters, calibration and practical aspects of Hp(3) monitoring. A 2015 review

    International Nuclear Information System (INIS)

    Carinou, Eleftheria; Ferrari, Paolo; Bjelac, Olivera Ciraj; Gingaume, Merce; Merce, Marta Sans; O’Connor, Una

    2015-01-01

    A thorough literature review about the current situation on the implementation of eye lens monitoring has been performed in order to provide recommendations regarding dosemeter types, calibration procedures and practical aspects of eye lens monitoring for interventional radiology personnel. Most relevant data and recommendations from about 100 papers have been analysed and classified in the following topics: challenges of today in eye lens monitoring; conversion coefficients, phantoms and calibration procedures for eye lens dose evaluation; correction factors and dosemeters for eye lens dose measurements; dosemeter position and influence of protective devices. The major findings of the review can be summarised as follows: the recommended operational quantity for the eye lens monitoring is H p (3). At present, several dosemeters are available for eye lens monitoring and calibration procedures are being developed. However, in practice, very often, alternative methods are used to assess the dose to the eye lens. A summary of correction factors found in the literature for the assessment of the eye lens dose is provided. These factors can give an estimation of the eye lens dose when alternative methods, such as the use of a whole body dosemeter, are used. A wide range of values is found, thus indicating the large uncertainty associated with these simplified methods. Reduction factors from most common protective devices obtained experimentally and using Monte Carlo calculations are presented. The paper concludes that the use of a dosemeter placed at collar level outside the lead apron can provide a useful first estimate of the eye lens exposure. However, for workplaces with estimated annual equivalent dose to the eye lens close to the dose limit, specific eye lens monitoring should be performed. Finally, training of the involved medical staff on the risks of ionising radiation for the eye lens and on the correct use of protective systems is strongly recommended. (review)

  13. Neutron dosimetry using activation of thermoluminescent CaSO 4

    Science.gov (United States)

    Azorín, Juan; Gutiérrez, Alicia

    1984-11-01

    Sulfur activation in calcium sulfate doped with dysprosium (CaSO 4:Dy) thermoluminescent powder, which is bound in pure sulfur, has been used to measure the fast neutron dose at the tangential beam port of a Triga Mark III reactor. After a post-irradiation time of 3 d, the dosimeters were annealed at 600°C for 30 min in order to erase all the thermoluminescence acquired during the irradiation. The dosimeters were then stored to allow self-irradiation by betas from 32P produced by sulfur activation. The thermoluminescent signal accumulated during a post-irradiation time of 20 d due to a neutron fluence of 2.2 × 10 11 n/cm 2 was equivalent to an absorbed dose of 10 mGy of 60Co gamma rays. The thermoluminescence as a function of fast neutron dose fitted to a straight line on a log-log scale from 1 Gy to 10 4Gy.

  14. Pilot study for natural radiation survey

    International Nuclear Information System (INIS)

    Brown, L.; Driscoll, C.M.H.; Green, B.M.R.; Miles, J.C.H.

    1983-01-01

    NRPB's national survey of natural radiation exposure in homes commenced in 1982 and will run until 1984. A pilot survey was undertaken in over 100 homes for one year, using passive thermoluminescent dosemeters to measure external radiation from terrestrial and cosmic sources and passive radon dosemeters to measure the radon-222 gas concentration. A preliminary analysis of the results obtained from the pilot survey is given. The main value of the pilot survey was in providing experience and various administrative and scientific procedures have been simplified or automated for the national survey. (U.K.)

  15. Dosimetry in a torso phantom during a mammography

    International Nuclear Information System (INIS)

    Hernandez O, M.; Duran M, H. A.; Pinedo S, A.; Salas L, M. A.; Hernandez D, V. M.; Vega C, H. R.; Rivera M, T.; Ventura M, J.

    2009-10-01

    Two dosimetric magnitudes, the absorbed dose and the kerma in air to the entrance of torso have been determined. These dosimetric magnitudes are due to the radiation that is dispersed in the mammary gland when the patient undergoes a mammography study. The kerma to the entrance of the torso and the absorbed dose by the torso was obtained in a phantom of paraffin and with thermoluminescent dosemeters of ZrO 2 . The dosemeters were placed on the surface of the torso phantom while the mammography was carried out. (author)

  16. Gamma dosimetry of high doses

    International Nuclear Information System (INIS)

    Martinez C, T.; Galvan G, A.; Canizal, G.

    1991-01-01

    The gamma dosimetry of high doses is problematic in almost all the classic dosemeters either based on the thermoluminescence, electric, chemical properties, etc., because they are saturated to very high dose and they are no longer useful. This work carries out an investigation in the interval of high doses. The solid system of heptahydrate ferrous sulfate, can be used as solid dosemeter of routine for high doses of radiation. The proposed method is simple, cheap and it doesn't require sophisticated spectrophotometers or spectrometers but expensive and not common in some laboratories

  17. Chronology of Chichen Izta, evidences by thermoluminescence

    International Nuclear Information System (INIS)

    Ramirez, A.; Schaaf, P.; Gonzalez, P.R.; Chung, H.

    2002-01-01

    The establishment of the chronology of Chichen Itza is argued by its stylistic evocations, architectural and iconographical as well as its history related in the sources, as the Chilam Balam, where is named the arrival of foreign tribes as the Xiu and the Itzaes and its settlement in Chichen Itza as well as the return to its origin place. The material evidence of such transition is reflected in the ceramic change of the Pizarra Puuc to the Pizarra Chichen, however, in 13 stratigraphic wells realized in different points of Chichen Itza, it was not found evidence of the layer of the Pizarra Puuc period. In this work the obtained ages of dating by thermoluminescence (TL) are presented, of four samples of ceramics found in the different wells. When our results are compared with the estimated ages for the different groups of ceramics, it was observed that none of they corresponded to the Pizarra Puuc period. These results strenghten the observations realized during excavations, causing the suggestion of new inferences related to the Chichen Itza boom and its relationship with the Maya Classic Period. The used technique for the TL dating was that of fine grain. The interval of artificial dose, of beta radiation of Strontium 90 was between 2 and 30 Gy. The TL signal was obtained heating the samples at 10 C/s until reaching 500 C in the TL Daybreak reader equipment. In order to know the annual dose rate of concentration of potassium ( 40 K) by means of the microanalysis technique in the scanning electron microscope was determined, while the uranium content (2 38 U) and thorium ( 232 Th) were determined by means of the neutron activation analysis technique using the TRIGA Mark III nuclear reactor. The contribution of the gamma radiation of the soil as well as the contribution of the cosmic radiation was measured with TL dosemeters of CaSO 4 :Dy + Ptfe once the paleodose was known and the annual dose rate of each sample of its age was estimated. (Author)

  18. Glass formation, magnetic properties and magnetocaloric effect of ternary Ho–Al–Co bulk metallic glass

    International Nuclear Information System (INIS)

    Zhang, Huiyan; Li, Ran; Ji, Yunfei; Liu, Fanmao; Luo, Qiang; Zhang, Tao

    2012-01-01

    A ternary Ho–Al–Co system with high glass-forming ability (GFA) was developed and fully glassy rods with diameters up to 1 cm can be produced for the best glass former of Ho 55 Al 27.5 Co 17.5 alloy. The thermal stability and low-temperature magnetic properties of the Ho 55 Al 27.5 Co 17.5 bulk metallic glass (BMG) were studied. The magnetic transition temperature of this alloy is ∼14 K as determined by the thermomagnetic measurement. Two indicators, i.e. isothermal magnetic entropy change (ΔS M ) and the relative cooling power (RCP), were adopted to evaluate the magnetocaloric effect (MCE) of the alloy under a low magnetic field up to 2 T, which can be generated by permanent magnets. The values of |ΔS M | and RCP are 7.98 J kg −1 K −1 and 191.5 J kg −1 , respectively. The Ho 55 Al 27.5 Co 17.5 BMG with good MCE and high GFA provides an attractive candidate for magnetic refrigeration applications, like hydrogen liquefaction and storage. - Highlights: ► A ternary Ho–Al–Co BMG system with high glass-forming ability was developed. ► Fully glassy rods of Ho 55 Al 27.5 Co 17.5 alloy were produced up to 1 cm in diameter. ► The thermal stability and magnetic properties of the BMG were evaluated. ► The BMG exhibits good magnetocaloric effect under a low magnetic field up to 2 T.

  19. Thermoluminescence on ZrO{sub 2} films with different dopants; Termoluminiscencia en peliculas de ZrO{sub 2} con distintos impurificantes

    Energy Technology Data Exchange (ETDEWEB)

    Ceron R, P. V.; Rivera M, T.; Ramos G, A. I.; Guzman M, J.; Montes R, E., E-mail: victceronr@hotmail.com [IPN, Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Av. Legaria No. 694, Col. Irrigacion, 11500 Mexico D. F. (Mexico)

    2015-10-15

    Full text: The metal oxides doped with rare earths have presented good thermoluminescent properties for certain wavelengths in the UV. With respect to zirconium oxide exist several studies in which were incorporated impurities and their properties as dosimeter in several regions of the electromagnetic spectrum were analyzed. Because of this background, in this material thermoluminescent glow curves induced by UV in films of ZrO{sub 2}:Eu and ZrO{sub 2}:Tb were studied for comparison with the response of the material doped with two rare earths (ZrO{sub 2}:Eu + Tb). Samples were deposited on glass by ultrasonic spray pyrolysis technique with different synthesis parameters. It was found that the strongest Tl response was to ZrO{sub 2} film doped with terbium (14 times more intense than the film of ZrO{sub 2}:Eu and 6 times the response of ZrO{sub 2}:Eu + Tb). (Author)

  20. Calibration of PADC-based neutron area dosemeters in the neutron field produced in the treatment room of a medical LINAC

    International Nuclear Information System (INIS)

    Bedogni, R.; Domingo, C.; Esposito, A.; Gentile, A.; García-Fusté, M.J.; San-Pedro, M. de; Tana, L.; D’Errico, F.; Ciolini, R.; Di Fulvio, A.

    2013-01-01

    PADC-based nuclear track detectors have been widely used as convenient ambient dosemeters in many working places. However, due to the large energy dependence of their response in terms of ambient dose equivalent (H ∗ (10)) and to the diversity of workplace fields in terms of energy distribution, the appropriate calibration of these dosemeters is a delicate task. These are among the reasons why ISO has introduced the 12789 Series of Standards, where the simulated workplace neutron fields are introduced and their use to calibrate neutron dosemeters is recommended. This approach was applied in the present work to the UAB PADC-based nuclear track detectors. As a suitable workplace, the treatment room of a 15 MV Varian CLINAC DHX medical accelerator, located in the Ospedale S. Chiara (Pisa), was chosen. Here the neutron spectra in two points of tests (1.5 m and 2 m from the isocenter) were determined with the INFN-LNF Bonner Sphere Spectrometer equipped with Dysprosium activation foils (Dy-BSS), and the values of H ∗ (10) were derived on this basis. The PADC dosemeters were exposed in these points. Their workplace specific H*(10) responses were determined and compared with those previously obtained in different simulated workplace or reference (ISO 8529) neutron fields. - Highlights: ► The neutron field of a medical LINAC was used to calibrate PADC neutron dosemeters. ► The neutron spectra were derived with a Dy-foil based Bonner Sphere Spectrometer. ► Workplace specific calibration factor were derived for the PADC dosemeters. ► These factors were compared with those obtained in reference neutron fields