  1. Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine

    Energy Technology Data Exchange (ETDEWEB)

    Yiannakkaras, C; Papadopoulos, N; Christodoulides, G [Department of Medical Physics, Nicosia General Hospital, 1450 Nicosia (Cyprus)

    1999-12-31

    The performance and beam characteristics of the new PANTAK THERAPAX HF225 X-ray therapy machine have been measured, evaluated and discussed. Eight beam qualities within the working range of generating potentials between 50 and 225 kVp are used in our department. These beam qualities have been investigated in order to provide a data base specific to our machine. Beam Quality, Central Axis Depth Dose, Output, Relative Field Uniformity and Timer Error were investigated. (authors) 11 refs., 4 figs., 9 tabs.

  2. Philips high tension generator (x-ray machine) testing for baby ebm (electron beam machine) project

    International Nuclear Information System (INIS)

    Norman Awalludin; Leo Kwee Wah; Abu Bakar Mhd Ghazali

    2005-01-01

    This paper describes the test of the HT system (from X-ray machine) for usage of the mini EBM (Electron Beam Machine). It consists the procedures of the installation, the safety procedures when deals with HT, modification of the system for testing purpose and the technique/method for testing the HT system. As a result, the voltage for the HT system and the electron gun (filament) current can be measured. Based on the results, suitability of the machine for baby EBM could be confirmed. (Author)

  3. Characteristics determination of Tanka X-ray Diagnostic Machine Model RTO-125

    International Nuclear Information System (INIS)

    Trijoko, Susetyo; Nasukha; Suyati; Nugroho, Agung.

    1993-01-01

    Characteristics determination of Tanka X-ray diagnostic machine model RTO-125. The characteristics of X-ray machine used for examining patient should be known. The characteristics studied in this paper include : X-ray beam profile, coincidence of the light field with radiation field, peak voltage, radiation quality, stability of exposures, and linearity of exposures against time. Beam profile and radiation-field alignment were determined using X-ray film. Winconsin kVp test cassette was used to measure peak voltage. The quality of the radiation, represented by half-value layer (HVL), was measured using aluminium step-wedge. Stability and linearity of exposures were measured using ionization chamber detector having an air volume of 40 cc. The results of this study were documented for the TANKA X-ray machine model RTO-125 of PSPKR BATAN, and the method of this study could be applied for X-ray diagnostic machine in general. (authors). 6 refs., 2 tabs., 6 figs

  4. Accelerators for E-beam and X-ray processing

    Energy Technology Data Exchange (ETDEWEB)

    Auslender, V.L. E-mail: auslen@inp.nsk.su; Bryazgin, A.A.; Faktorovich, B.L.; Gorbunov, V.A.; Kokin, E.N.; Korobeinikov, M.V.; Krainov, G.S.; Lukin, A.N.; Maximov, S.A.; Nekhaev, V.E.; Panfilov, A.D.; Radchenko, V.N.; Tkachenko, V.O.; Tuvik, A.A.; Voronin, L.A

    2002-03-01

    During last years the demand for pasteurization and desinsection of various food products (meat, chicken, sea products, vegetables, fruits, etc.) had increased. The treatment of these products in industrial scale requires the usage of powerful electron accelerators with energy 5-10 MeV and beam power at least 50 kW or more. The report describes the ILU accelerators with energy range up to 10 MeV and beam power up to 150 kW.The different irradiation schemes in electron beam and X-ray modes for various products are described. The design of the X-ray converter and 90 deg. beam bending system are also given.

  5. The effect of target thickness on x-ray production by FXR [Flash X-Ray Machine

    International Nuclear Information System (INIS)

    Back, N.L.

    1986-01-01

    The electron-photon transport code SANDYL has been used to calculate the x-ray flux for a simplified Flash X-Ray Machine (FXR) bullnose geometry. Four different thicknesses (24.5, 36.75, 49, and 61.25 mils) were used for the tantalum bremsstrahlung target in order to study the effect of target thickness on the FXR output. The calculations were performed for a parallel 17 MeV electron beam, and the resulting angular distributions were then used to compute the forward flux for the more realistic case of a converging beam. Over the range of thicknesses studied, the x-ray energy content per steradian on axis was essentially independent of target thickness. The main reason for this is that, while the total x-ray flux coming out of the target increases with increasing target thickness, the angular width of that flux also increases. The implications for target wheel design are discussed. 3 refs., 7 figs

  6. Abreu system - A dosimetric system to evaluate basic parameters of photofluorographic X-ray machine

    International Nuclear Information System (INIS)

    Feital, J.C.

    1987-01-01

    In Brazil, photofluorographic X-ray machines are used for cuberculosis mass screening throughout the country. The exact number of these X-ray equipment is unknown but it is estimated to be around 1000 operating units. Twelve million miniature chest radiographs are taken per year. In order to make local inspections speedier and also aiming at its postal use, a system has been developed wich evaluates the entrace exposure of the patient, the X-ray beam half-value layer ( leading to the evaluation of the tube's total filtration ) and the beam's field size. It consists of a piece of cardboard where filters, TLDs and X-ray films are inserted. So far the system has been tested in 53 X-ray machines in Rio de Janeiro. The results show that it can be used in a national survey program. (Author) [pt

  7. Heat transfer issues in high-heat-load synchrotron x-ray beams

    International Nuclear Information System (INIS)

    Khounsary, A.M.; Mills, D.M.

    1994-09-01

    In this paper, a short description of the synchrotron radiation x-ray sources and the associated power loads is given, followed by a brief description of typical synchrotron components and their heat load. It is emphasized that the design goals for most of these components is to limit (a) temperature, (b) stresses, or (c) strains in the system. Each design calls for a different geometry, material selection, and cooling scheme. Cooling schemes that have been utilized so far are primarily single phase and include simple macrochannel cooling, microchannel cooling, contact cooling, pin-post cooling, porous-flow cooling, jet cooling, etc. Water, liquid metals, and various cryogenic coolants have been used. Because the trend in x-ray beam development is towards brighter (i.e., more powerful) beams and assuming that no radical changes in the design of x-ray generating machines occurs in the next few years, it is fair to state that the utilization of various effective cooling schemes and, in particular, two-phase flow (e.g., subcooled boiling) warrants further investigation. This, however, requires a thorough examination of stability and reliability of two-phase flows for high-heat-flux components operating in ultrahigh vacuum with stringent reliability requirements

  8. Vertical beam size measurement in the CESR-TA e{sup +}e{sup −} storage ring using x-rays from synchrotron radiation

    Energy Technology Data Exchange (ETDEWEB)

    Alexander, J.P.; Chatterjee, A.; Conolly, C.; Edwards, E.; Ehrlichman, M.P.; Fontes, E. [Cornell University, Ithaca, NY 14853 (United States); Heltsley, B.K., E-mail: bkh2@cornell.edu [Cornell University, Ithaca, NY 14853 (United States); Hopkins, W.; Lyndaker, A.; Peterson, D.P.; Rider, N.T.; Rubin, D.L.; Savino, J.; Seeley, R.; Shanks, J. [Cornell University, Ithaca, NY 14853 (United States); Flanagan, J.W. [High Energy Accelerator Research Organization (KEK), Tsukuba (Japan)

    2014-06-01

    We describe the construction and operation of an X-ray beam size monitor (xBSM), a device measuring e{sup +} and e{sup −} beam sizes in the CESR-TA storage ring using synchrotron radiation. The device can measure vertical beam sizes of 10–100μm on a turn-by-turn, bunch-by-bunch basis at e{sup ±} beam energies of ∼2GeV. At such beam energies the xBSM images X-rays of ϵ≈1–10keV (λ≈0.1–1nm) that emerge from a hard-bend magnet through a single- or multiple-slit (coded aperture) optical element onto an array of 32 InGaAs photodiodes with 50μm pitch. Beamlines and detectors are entirely in-vacuum, enabling single-shot beam size measurement down to below 0.1 mA (2.5×10{sup 9} particles) per bunch and inter-bunch spacing of as little as 4 ns. At E{sub b}=2.1GeV, systematic precision of ∼1μm is achieved for a beam size of ∼12μm; this is expected to scale as ∝1/σ{sub b} and ∝1/E{sub b}. Achieving this precision requires comprehensive alignment and calibration of the detector, optical elements, and X-ray beam. Data from the xBSM have been used to extract characteristics of beam oscillations on long and short timescales, and to make detailed studies of low-emittance tuning, intra-beam scattering, electron cloud effects, and multi-bunch instabilities.

  9. X-ray beam generator

    International Nuclear Information System (INIS)

    Koller, T.J.; Randmer, J.A.

    1977-01-01

    A method of minimizing the preferential angular absorption of the divergent beam from an X-ray generator is described. The generator consists of an X-ray shielded housing with an X-ray transmissive window symmetrically placed in radial alignment with a focal spot area on a sloped target surface of an X-ray tube in the housing. The X-ray tube may be of the stationary anode type or of the rotating anode type. (U.K.)

  10. Developement of EBM from x-ray machine

    International Nuclear Information System (INIS)

    Muhamad Zahidee Taat; Leo Kwee Wah; Mohd Rizal Md Chulan; Abu Bakar Ghazali; Razali Hamzah; Abu Bakar Harun; Wan Hadi Wan Abu Bakar; Mohd Rizal Mamat

    2004-01-01

    This paper explains the works being done to convert an old x-ray machine into a low energy electron beam accelerator. In this project, we replace the x-ray tube from the machine with an electron accelerating tube while maintaining the high voltage power supply of 160 kV. Components involved in this project development are understanding the physics of accelerated particles, designing the ultra high vacuum system of at least 10-6 torr to vacuum the tube, assembling the high voltage supply and electron gun to accelerator tube and installing a thin titanium at the end of the tube as a window for the electrons to escape to atmosphere. Two stages of pumping are needed to meet the vacuum requirement; the first is for rough pumping down to 10 -3 torr and then subsequently switching to the second stage that continues pumping to reach the required level. For the first stage, we use a rotary pump and for the later either a turbo molecular pump or a diffusion pump. Furthermore, it is also essential to design and fabricate a connector with suitable insulation material to connect the high tension (HT) cable from the x-ray machine with the cathode of the accelerating tube. The electrons that escape from the window into atmosphere are collected and measured using a faraday cup meter. We expect to get some initial data from this setup before the seminar and hopefully will present it together with the theoretical calculation and design of the main components. (Author)

  11. Measurement of X-ray beam emittance using crystal optics at an X-ray undulator beamline

    CERN Document Server

    Kohmura, Y; Awaji, M; Tanaka, T; Hara, T; Goto, S; Ishikawa, T

    2000-01-01

    We present a method of using crystal optics to measure the emittance of the X-ray source. Two perfect crystals set in (++) configuration work as a high-resolution collimator. The phase-space diagram (i.e. beam cross-section and angular distribution) could be determined without any assumptions on the light source. When the measurement is done at short wavelength radiation from undulator, the electron beam emittance is larger than the diffraction limit of the X-rays. Therefore, the electron beam emittance could be estimated. The measurement was done with the hard X-rays of 18.5 and 55 keV from an undulator beamline, BL 47XU, of SPring-8. The horizontal emittance of the X-ray beam was estimated to be about 7.6 nmrad, close to the designed electron beam emittance of the storage ring (7 nmrad). Some portions of the instrumental functions, such as the scattering by filters and windows along the beamline and the slight bent of the crystal planes of the monochromator, could not be precisely evaluated, but an upper li...

  12. Measurement of X-ray beam emittance using crystal optics at an X-ray undulator beamline

    International Nuclear Information System (INIS)

    Kohmura, Yoshiki; Suzuki, Yoshio; Awaji, Mitsuhiro; Tanaka, Takashi; Hara, Toru; Goto, Shunji; Ishikawa, Tetsuya

    2000-01-01

    We present a method of using crystal optics to measure the emittance of the X-ray source. Two perfect crystals set in (++) configuration work as a high-resolution collimator. The phase-space diagram (i.e. beam cross-section and angular distribution) could be determined without any assumptions on the light source. When the measurement is done at short wavelength radiation from undulator, the electron beam emittance is larger than the diffraction limit of the X-rays. Therefore, the electron beam emittance could be estimated. The measurement was done with the hard X-rays of 18.5 and 55 keV from an undulator beamline, BL 47XU, of SPring-8. The horizontal emittance of the X-ray beam was estimated to be about 7.6 nmrad, close to the designed electron beam emittance of the storage ring (7 nmrad). Some portions of the instrumental functions, such as the scattering by filters and windows along the beamline and the slight bent of the crystal planes of the monochromator, could not be precisely evaluated, but an upper limit for the vertical emittance of the electron beam could be obtained as 0.14 nmrad

  13. Hard X-ray emission spectroscopy with pink beam

    Energy Technology Data Exchange (ETDEWEB)

    Kvashnina, Kristina O.; Rossberg, Andre; Exner, Joerg; Scheinost, Andreas C. [Helmholtz-Zentrum Dresden-Rossendorf e.V., Dresden (Germany). Molecular Structures

    2017-06-01

    Valence-band X-ray emission spectroscopy (XES) with a ''pink beam'', i.e. a beam with large energy bandwidth produced by a double-multilayer monochromator, is introduced here to overcome the weak count rate of monochromatic beams produced by conventional double-crystal monochromators. Our results demonstrate that - in spite of the large bandwidth in the order of 100 eV - the high spectral resolution of the Johann-type spectrometer is maintained, while the two orders of magnitude higher flux greatly reduces the required counting time. The short working distance Johann-type X-ray emission spectrometer and multilayer monochromator is available at ROBL.

  14. Panoramic dental X-ray machine X-motion drive

    International Nuclear Information System (INIS)

    Cushman, R.H.; Flynn, J.J.

    1980-01-01

    A panoramic dental x-ray machine is described which provides continuous and discontinuous radiographic images of the dental arch area of a patient. The systems for moving the chair and the column which carries the x-ray source and camera are specified. (U.K.)

  15. The quality of high-energy X-ray beams

    International Nuclear Information System (INIS)

    LaRiviere, P.D.

    1989-01-01

    Supplement 17 of the British Journal of Radiology is a survey of central-axis depth doses for radiotherapy machines, patterned largely on BJR Supplement 11 (1972). Inspection of high-energy X-ray depth doses for a 10 x 10 cm field at an SSD of 100 cm disclosed large differences between the two sets of data, especially for qualities above 8 MV, e.g. a depth dose of 80% at 10 cm is rated at about 19 MV according to BJR Supplement 11, and 23 MV according to BJR Supplement 17. It was found that Supplement 17 depth-dose data above 8 MV were erratic, but Supplement 11 data could be represented by an analytical expression, providing a unique means of assigning MV quality. It was also found that dose-weighted average energy of the filtered beam plotted smoothly against depth dose. For dosimetric purposes, it is suggested that this parameter be used as a true measure of beam quality, removing discrepancies introduced by the use of nominal MV for this purpose. (author)

  16. Dose inspection and risk assessment on radiation safety for the use of non-medical X-ray machines in Taiwan

    Science.gov (United States)

    Hsu, Fang-Yuh; Hsu, Shih-Ming; Chao, Jiunn-Hsing

    2017-11-01

    The subject of this study is the on-site visits and inspections of facilities commissioned by the Atomic Energy Council (AEC) in Taiwan. This research was conducted to evaluate the possible dose and dose rate of cabinet-type X-ray equipment with nominal voltages of 30-150 kV and open-beam (portable or handheld) equipment, taking both normal operation and possibly abnormal operation conditions into account. Doses and dose rates were measured using a plastic scintillation survey meter and an electronic personal dosimeter. In total, 401 X-ray machines were inspected, including 139 units with nominal voltages of 30-50 kV X-ray equipment, 140 units with nominal voltages of 50-150 kV, and 122 open-beam (portable or handheld) X-ray equipment. The investigated doses for radiation workers and non-radiation workers operating cabinet-type X-ray equipment under normal safety conditions were all at the background dose level. Several investigated dose rates at the position of 10 cm away from the surface of open-beam (portable or handheld) X-ray equipment were very high, such X-ray machines are used by aeronautical police for the detection of suspected explosives, radiation workers are far away (at least 10 m away) from the X-ray machine during its operation. The doses per operation in X-ray equipment with a 30-50 kV nominal voltage were less than 1 mSv in all cases of abnormal use. Some doses were higher than 1 mSv per operation for X-ray equipment of 50-150 kV nominal voltage X-ray. The maximum dose rates at the beam exit have a very wide range, mostly less than 100 μSv/s and the largest value is about 3.92 mSv/s for open-beam (portable or handheld) X-ray devices. The risk induced by operating X-ray devices with nominal voltages of 30-50 kV is extremely low. The 11.5 mSv dose due to one operation at nominal voltage of 50-150 kV X-ray device is equivalent to the exposure of taking 575 chest X-rays. In the abnormal use of open-beam (portable or handheld) X-ray equipment, the

  17. Quality control procedure for a general diagnostic x-ray machine

    International Nuclear Information System (INIS)

    Md Saion Salikin; Mazlyfarina Mohamad

    2001-01-01

    Performance tests of a particular medical diagnostic x-ray machine have to be carried out regularly in order to ensure that the machine always complies with the required standard. A performance report which is prepared based on the performance tests on the x-ray machine is used as one of the requirement by the authority for issuance and renewal licence to operate and use of the x-ray machine in a clinic or hospital. The Ministry of Health will only issue a licence to the clinic or hospital to use and operate x-ray machines, if the machines have complied with the standards. The clinic or hospital may get the service to carry out performance tests on a diagnostic x-ray machine from any licence H holder, issued by the authority under Atomic Energy Licensing Act 1984. A comparative study between the standard procedure and other procedures on performance test of a general diagnostic x-ray machine is carried out and presented in brief in this paper. The criteria for compliance or otherwise as required by the Ministry of Health Malaysia is discussed and explained in brief. (Author)

  18. X-ray beam qualities for dental radiology purposes

    International Nuclear Information System (INIS)

    Santos, Marcus Aurelio P. dos; Fragoso, Maria da Conceicao de F.; Lima, Ricardo de A.; Hazim, Clovis A.

    2009-01-01

    In order to establish characteristics or properties of equipment for diagnostic radiology, e.g. ion chambers and semiconductor detectors, calibration laboratories offer a set of well-defined radiation conditions, called X-ray qualities, which can be used for many Physics studies and medical purposes. The standardization of radiation qualities has been carried out in several fields of study, but little attention has been given to the area of dental radiology, mainly for medical and physical applications using single-phase units with half-wave rectification. For this reason, a single-phase dental unit with adjustable peak voltage and tube current, called 'variable potential X-ray equipment', was developed aiming to define X-ray beam qualities for test and calibrations purposes. X-ray spectra at 50, 60 and 70 kVp were determined by using a CdTe detector and compared with those obtained for ten commercial X-ray dental units. As a result of this study, a set of X-ray qualities for the variable potential X-ray equipment was determined. The X-ray qualities spectra were utilized as reference for determination of a new set of X-ray qualities characterized for a constant potential X-ray equipment. Thus, sets of X-ray qualities were standardized and implemented in two X-ray laboratories: one with the variable potential X-ray equipment and other with constant potential X-ray equipment. These reference X-ray beam qualities should be used for test and calibration purposes involving scientific studies and services. (author)

  19. Surveillance of X-ray machines in Israel

    International Nuclear Information System (INIS)

    Donagi, A.; Hai, J.; Kuszpet, M.

    1980-01-01

    A nationwide surveillance of X-ray machines is carried out in Israel by the Research Institute for Environmental Health, Ministry of Health. At present, diagnostic X-ray machines are surveyed at least once every two years, while dental machines are surveyed once every five years. The investigated parameters include measurement of output, scattered radiation, X-ray-light-field alignment, HVL, inherent filtration, structural shielding, etc. In order to compare X-ray techniques used in different hospitals in Israel, the NEXT (Nationwide Evaluation of X-ray Trends) program, which was developed by the BRH (Bureau of Radiological Health, USPHS) was utilized. On the basis of the findings of this project, necessary correction steps were taken in order to reduce the radiation doses to both personnel and patients. Further activities of the Institute include the estimation of doses delivered to pregnant women who were irradiated during the gestation period. This information is sent to a professional committee, which decides whether or not to perform an abortion. Recently, the new recommendations of ICRP 26 (International Commission on Radiological Protection) were implemented in Israel. Therefore, at present doses over 150 mR/month are reported to the Institute by the Soreq Personal Dosimetry Service, and the causes of this exposure are investigated. (author)

  20. A Study of the Resolution of Dental Intraoral X-Ray Machines

    International Nuclear Information System (INIS)

    Kim, Seon Ju; Chung, Hyon De

    1990-01-01

    The purpose of this study was to assess the resolution and focal spot size of dental X-ray machines. Fifty dental X-ray machines were selected for measuring resolution and focal spot size. These machines were used in general dental clinics. The time on installation of the X-ray machine varies from 1 years to 10 years. The resolution of these machines was measured with the test pattern. The focal spot size of these machines was measured with the star test pattern. The following results were obtained: 1. The resolution of dental intraoral X-ray machines was not significantly changed in ten years. 2. The focal spot size of dental intraoral X-ray machines was not significantly increased in ten years. The statistical analysis between the mean focal spot size and nominal focal spot size was significant at the 0.05 level about the more than 3 years used machines.

  1. Using a tandem ionization chamber for quality control of X-ray beams

    International Nuclear Information System (INIS)

    Yoshizumi, Maira T.; Caldas, Linda V.E.

    2011-01-01

    X-ray beam qualities are defined by both the mean energies and by the half-value layers (HVL). Many international protocols use the half-value layer and the beam voltage to characterize the X-ray beam quality. A quality control program for X-ray equipment includes the constancy check of beam qualities, i.e., the periodical verification of the half-value layer, which can be a time consumable procedure. A tandem ionization chamber, developed at Instituto de Pesquisas Energeticas e Nucleares, was used to determine the HVL and its constancy for five radiotherapy standard beam qualities. This ionization chamber is composed by two sensitive volumes with inner electrodes made of different materials: aluminum and graphite. The beam quality constancy check test was performed during two months and the maximum variation obtained was 1.24% for the radiation beam quality T-10. This result is very satisfactory according to national recommendations. (author)

  2. Dose properties of x-ray beams produced by laser-wakefield-accelerated electrons

    International Nuclear Information System (INIS)

    Kainz, K K; Hogstrom, K R; Antolak, J A; Almond, P R; Bloch, C D

    2005-01-01

    Given that laser wakefield acceleration (LWFA) has been demonstrated experimentally to accelerate electron beams to energies beyond 25 MeV, it is reasonable to assess the ability of existing LWFA technology to compete with conventional radiofrequency linear accelerators in producing electron and x-ray beams for external-beam radiotherapy. We present calculations of the dose distributions (off-axis dose profiles and central-axis depth dose) and dose rates of x-ray beams that can be produced from electron beams that are generated using state-of-the-art LWFA. Subsets of an LWFA electron energy distribution were propagated through the treatment head elements (presuming an existing design for an x-ray production target and flattening filter) implemented within the EGSnrc Monte Carlo code. Three x-ray energy configurations (6 MV, 10 MV and 18 MV) were studied, and the energy width ΔE of the electron-beam subsets varied from 0.5 MeV to 12.5 MeV. As ΔE increased from 0.5 MeV to 4.5 MeV, we found that the off-axis and central-axis dose profiles for x-rays were minimally affected (to within about 3%), a result slightly different from prior calculations of electron beams broadened by scattering foils. For ΔE of the order of 12 MeV, the effect on the off-axis profile was of the order of 10%, but the central-axis depth dose was affected by less than 2% for depths in excess of about 5 cm beyond d max . Although increasing ΔE beyond 6.5 MeV increased the dose rate at d max by more than 10 times, the absolute dose rates were about 3 orders of magnitude below those observed for LWFA-based electron beams at comparable energies. For a practical LWFA-based x-ray device, the beam current must be increased by about 4-5 orders of magnitude. (note)

  3. Absorbed and effective dose from periapical radiography by portable intraoral x-ray machine

    International Nuclear Information System (INIS)

    Cho, Jeong Yeon; Han, Won Jeong; Kim, Eun Kyung

    2007-01-01

    The purpose of this study was to measure the absorbed dose and to calculate the effective dose for periapical radiography done by portable intraoral x-ray machines. 14 full mouth, upper posterior and lower posterior periapical radiographs were taken by wall-type 1 and portable type 3 intraoral x-ray machines. Thermoluminescent dosemeters were placed at 23 sites at the layers of the tissue-equivalent ART woman phantom for dosimetry. Average tissue absorbed dose and radiation weighted dose were calculated for each major anatomical site. Effective dose was calculated using 2005 ICRP tissue weighted factors. On 14 full mouth periapical radiographs, the effective dose for wall-type x-ray machine was 30 Sv; for portable x-ray machines were 30 Sv, 22 Sv, 36 Sv. On upper posterior radiograph, the effective dose for wall-type x-ray machine was 4 Sv; for portable x-ray machines doses were 4 Sv, 3 Sv, 5 Sv. On lower posterior radiograph, the effective dose for wall type x-ray machine was 5 Sv; for portable x-ray machines doses were 4 Sv, 4 Sv, 5 Sv. Effective doses for periapical radiographs performed by portable intraoral x-ray machines were similar to doses for periapical radiographs taken by wall type intraoral x-ray machines

  4. Advances in kilovoltage x-ray beam dosimetry

    Science.gov (United States)

    Hill, Robin; Healy, Brendan; Holloway, Lois; Kuncic, Zdenka; Thwaites, David; Baldock, Clive

    2014-03-01

    This topical review provides an up-to-date overview of the theoretical and practical aspects of therapeutic kilovoltage x-ray beam dosimetry. Kilovoltage x-ray beams have the property that the maximum dose occurs very close to the surface and thus, they are predominantly used in the treatment of skin cancers but also have applications for the treatment of other cancers. In addition, kilovoltage x-ray beams are used in intra operative units, within animal irradiators and in on-board imagers on linear accelerators and kilovoltage dosimetry is important in these applications as well. This review covers both reference and relative dosimetry of kilovoltage x-ray beams and provides recommendations for clinical measurements based on the literature to date. In particular, practical aspects for the selection of dosimeter and phantom material are reviewed to provide suitable advice for medical physicists. An overview is also presented of dosimeters other than ionization chambers which can be used for both relative and in vivo dosimetry. Finally, issues related to the treatment planning and the use of Monte Carlo codes for solving radiation transport problems in kilovoltage x-ray beams are presented.

  5. Nanoparticle-Assisted Scanning Focusing X-Ray Therapy with Needle Beam X Rays.

    Science.gov (United States)

    Davidson, R Andrew; Guo, Ting

    2016-01-01

    In this work, we show a new therapeutic approach using 40-120 keV X rays to deliver a radiation dose at the isocenter located many centimeters below the skin surface several hundred times greater than at the skin and how this dose enhancement can be augmented with nanomaterials to create several thousand-fold total dose enhancement effect. This novel approach employs a needle X-ray beam directed at the isocenter centimeters deep in the body while continuously scanning the beam to cover a large solid angle without overlapping at the skin. A Monte Carlo method was developed to simulate an X-ray dose delivered to the isocenter filled with X-ray absorbing and catalytic nanoparticles in a water phantom. An experimental apparatus consisting of a moving plastic phantom irradiated with a stationary 1 mm needle X-ray beam was built to test the theoretical predictions. X-ray films were used to characterize the dose profiles of the scanning X-ray apparatus. Through this work, it was determined that the X-ray dose delivered to the isocenter in a treatment voxel (t-voxel) underneath a 5 cm deep high-density polyethylene (HDPE) phantom was 295 ± 48 times greater than the surface dose. This measured value was in good agreement with the theoretical predicted value of 339-fold. Adding X-ray-absorbing nanoparticles, catalytic nanoparticles or both into the t-voxel can further augment the dose enhancement. For example, we predicted that adding 1 weight percentage (wp) of gold into water could increase the effective dose delivered to the target by onefold. Dose enhancement using 1 mm X-ray beam could reach about 1,600-fold in the t-voxel when 7.5 wp of 88 nm diameter silica-covered gold nanoparticles were added, which we showed in a previously published study can create a dose enhancement of 5.5 ± 0.46-fold without scanning focusing enhancement. Based on the experimental data from that study, mixing 0.02 wp 2.5 nm diameter small tetrakis hydroxymethyl phosphonium chloride (THPC

  6. Half value layer determination of Al and Cu for different x-ray beams and establishing a relationship between them

    International Nuclear Information System (INIS)

    Kharita, M. H.; Takeyeddin, M.; Wali, K.; Zahili, M.

    2005-03-01

    The Half Value Layer (HVL) of copper and aluminium for special standard x-rays beams (emitted from x-ray machine type Philips MGC30 used in Secondary Standard Dosimetry Laboratory (SSDL) have been determined. The relationship between HVL values of aluminium and copper was established. This relationship can be represented by a non-linear fit. These results are valuable for x-ray workers especially those working in the field of radiation calibration for beam energy determination. Using the results of this study, unavailable thickness of one metal can be replaced by the corresponding thickness of the other metal, for filtration and measurement of x-ray beam. (Authors)

  7. X-ray beam-position feedback system with easy-to-use beam-position monitor.

    Science.gov (United States)

    Park, Jae Yeon; Kim, Yesul; Lee, Sangsul; Lim, Jun

    2018-05-01

    X-ray beam-position stability is indispensable in cutting-edge experiments using synchrotron radiation. Here, for the first time, a beam-position feedback system is presented that utilizes an easy-to-use X-ray beam-position monitor incorporating a diamond-fluorescence screen. The acceptable range of the monitor is above 500 µm and the feedback system maintains the beam position within 3 µm. In addition to being inexpensive, the system has two key advantages: it works without a scale factor for position calibration, and it has no dependence on X-ray energy, X-ray intensity, beam size or beam shape.

  8. Shield device for controlling the dose of x-rays applied in an x-ray machine

    International Nuclear Information System (INIS)

    Charrier, P.

    1983-01-01

    This invention provides an improved shield for use with an x-ray machine. The shield can control the dose of x-rays applied by the machine in different areas without affecting the power of the x-rays. This is achieved with a shield especially designed and positioned to intercept with x-rays for longer or shorter periods in different areas during the taking of the picture, but not for the whole period of time necessary for taking this picture. Each area of the subject being x-rayed is exposed to full power x-rays. However, owing to the shield, the areas that require smaller dose receive these full power x-rays for a shorter portion of the time required to take the picture while the other areas that require larger dose of x-rays, receive the full power x-rays for a longer portion of the full period of time required to take the picture. To ensure this differential exposure, the shield is placed through the path of the x-rays and rotated about an axis which is generally transverse to the direction of travel of the x-rays to cut out some of said x-rays for different portions of the period of time necessary for taking the picture. The shield is preferably shaped to intercept x-rays for a longer period in some areas than in others depending on the required doses. A plurality of differently shaped shields can be provided to suit different picture taking situations

  9. Calibration of diagnostic x-ray machines using radiation exposure and radiographic parameters

    International Nuclear Information System (INIS)

    Agba, E.H.; Uloko, P. I.; Tyovenda, A. A.

    2011-01-01

    Calibration of diagnostic x-ray machines using radiation exposure and radiographic parameters of the x-ray machines has been carried out. Three phase diagnostic x-ray machines situated at Federal Medical Centre, Makurdi, General Hospital, Otukpo and Christian Hospital, Mkar were used for the calibration work. The radiation meter was used to measure x-ray radiation exposure. The result of this work demonstrates mR/mAs=C(KV p ) that there exist a power law relation of the form between the radiation exposure and the radiographic parameters of diagnostic x-ray machines, which can be used to estimate patient exposure during routine x-ray diagnostic examinations for wide range of operating parameters. The values of the power exponent n, constant c and total filtrations of the diagnostic x-ray machines have been estimated. These values for the diagnostic x-ray machines at the Federal Medical Centre, Makurdi are: 2.14, 0.88 and 2.77 respectively, for the one at the General Hospital, Otukpo are: 2.07, 0.76 and 2.68 respectively and that of the Christian Hospital, Mkar are: 2.01,0.69 and 2.61 respectively.

  10. BEAM OPTIMIZATION STUDY FOR AN X-RAY FEL OSCILLATOR AT THE LCLS-II

    Energy Technology Data Exchange (ETDEWEB)

    Qin, Weilun; Huang, S.; Liu, K.X.; Huang, Z; Ding, Y.; Maxwell, T.J.; Kim, K.-J.

    2016-06-01

    The 4 GeV LCLS-II superconducting linac with high repetition beam rate enables the possibility to drive an X-Ray FEL oscillator at harmonic frequencies *. Compared to the regular LCLS-II machine setup, the oscillator mode requires a much longer bunch length with a relatively lower current. Also a flat longitudinal phase space distribution is critical to maintain the FEL gain since the X-ray cavity has extremely narrow bandwidth. In this paper, we study the longitudinal phase space optimization including shaping the initial beam from the injector and optimizing the bunch compressor and dechirper parameters. We obtain a bunch with a flat energy chirp over 400 fs in the core part with current above 100 A. The optimization was based on LiTrack and Elegant simulations using LCLS-II beam parameters.

  11. Editorial: Focus on X-ray Beams with High Coherence

    Science.gov (United States)

    Robinson, Ian; Gruebel, Gerhard; Mochrie, Simon

    2010-03-01

    This editorial serves as the preface to a special issue of New Journal of Physics, which collects together solicited papers on a common subject, x-ray beams with high coherence. We summarize the issue's content, and explain why there is so much current interest both in the sources themselves and in the applications to the study of the structure of matter and its fluctuations (both spontaneous and driven). As this collection demonstrates, the field brings together accelerator physics in the design of new sources, particle physics in the design of detectors, and chemical and materials scientists who make use of the coherent beams produced. Focus on X-ray Beams with High Coherence Contents Femtosecond pulse x-ray imaging with a large field of view B Pfau, C M Günther, S Schaffert, R Mitzner, B Siemer, S Roling, H Zacharias, O Kutz, I Rudolph, R Treusch and S Eisebitt The FERMI@Elettra free-electron-laser source for coherent x-ray physics: photon properties, beam transport system and applications E Allaria, C Callegari, D Cocco, W M Fawley, M Kiskinova, C Masciovecchio and F Parmigiani Beyond simple exponential correlation functions and equilibrium dynamics in x-ray photon correlation spectroscopy Anders Madsen, Robert L Leheny, Hongyu Guo, Michael Sprung and Orsolya Czakkel The Coherent X-ray Imaging (CXI) instrument at the Linac Coherent Light Source (LCLS) Sébastien Boutet and Garth J Williams Dynamics and rheology under continuous shear flow studied by x-ray photon correlation spectroscopy Andrei Fluerasu, Pawel Kwasniewski, Chiara Caronna, Fanny Destremaut, Jean-Baptiste Salmon and Anders Madsen Exploration of crystal strains using coherent x-ray diffraction Wonsuk Cha, Sanghoon Song, Nak Cheon Jeong, Ross Harder, Kyung Byung Yoon, Ian K Robinson and Hyunjung Kim Coherence properties of the European XFEL G Geloni, E Saldin, L Samoylova, E Schneidmiller, H Sinn, Th Tschentscher and M Yurkov Fresnel coherent diffractive imaging: treatment and analysis of data G J

  12. Quantitative cone beam X-ray luminescence tomography/X-ray computed tomography imaging

    International Nuclear Information System (INIS)

    Chen, Dongmei; Zhu, Shouping; Chen, Xueli; Chao, Tiantian; Cao, Xu; Zhao, Fengjun; Huang, Liyu; Liang, Jimin

    2014-01-01

    X-ray luminescence tomography (XLT) is an imaging technology based on X-ray-excitable materials. The main purpose of this paper is to obtain quantitative luminescence concentration using the structural information of the X-ray computed tomography (XCT) in the hybrid cone beam XLT/XCT system. A multi-wavelength luminescence cone beam XLT method with the structural a priori information is presented to relieve the severe ill-posedness problem in the cone beam XLT. The nanophosphors and phantom experiments were undertaken to access the linear relationship of the system response. Then, an in vivo mouse experiment was conducted. The in vivo experimental results show that the recovered concentration error as low as 6.67% with the location error of 0.85 mm can be achieved. The results demonstrate that the proposed method can accurately recover the nanophosphor inclusion and realize the quantitative imaging

  13. Micro controller application as x-ray machine's high voltage controller

    International Nuclear Information System (INIS)

    Wiranto Budi Santoso; Beny Syawaludin

    2010-01-01

    The micro controller application as x-ray machine's high voltage controller has been carried out. The purpose of this micro controller application is to give an accurate high voltage supply to the x-ray tube so that the x ray machine could produce the result as expected. The micro controller based X-ray machine's high voltage controller receives an input voltage from the keypad. This input value is displayed in the LCD (Liquid Crystal Display) screen. Then micro controller uses this input data to drive the stepper motor. The stepper motor adjusts the high voltage auto transformer's output according to the input value. The micro controller is programmed using BASCOM-B051 compiler. The test results show that the stepper motor could rotate according to an input value. (author)

  14. X-ray emission as a diagnostic from pseudospark-sourced electron beams

    Energy Technology Data Exchange (ETDEWEB)

    Bowes, D., E-mail: david.bowes@strath.ac.uk [Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG (United Kingdom); Yin, H.; He, W.; Zhang, L.; Cross, A.W.; Ronald, K.; Phelps, A.D.R. [Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG (United Kingdom); Chen, D.; Zhang, P. [Computed Tomography Lab, School of Mathematical Sciences, Capital Normal University, Beijing 100048 (China); Chen, X.; Li, D. [Department of Electronic Engineering, Queen Mary University of London, London E1 4NS (United Kingdom)

    2014-09-15

    X-ray emission has been achieved using an electron beam generated by a pseudospark low-pressure discharge and utilised as a diagnostic for beam detection. A 300 A, 34 kV PS-sourced electron beam pulse of 3 mm diameter impacting on a 0.1 mm-thick molybdenum target generated X-rays which were detected via the use of a small, portable X-ray detector. Clear X-ray images of a micro-sized object were captured using an X-ray photodetector. This demonstrates the inducement of proton induced X-ray emission (PIXE) not only as an indicator of beam presence but also as a future X-ray source for small-spot X-ray imaging of materials.

  15. Method for reducing x-ray background signals from insertion device x-ray beam position monitors

    Directory of Open Access Journals (Sweden)

    Glenn Decker

    1999-11-01

    Full Text Available A method is described that provides a solution to the long-standing problem of stray radiation-induced signals on photoemission-based x-ray beam position monitors (BPMs located on insertion device x-ray beam lines. The method involves the introduction of a chicane into the accelerator lattice that directs unwanted x radiation away from the photosensitive x-ray BPM blades. This technique has been implemented at the Advanced Photon Source, and experimental confirmation of the technique is provided.

  16. Development of confocal micro X-ray fluorescence instrument using two X-ray beams

    International Nuclear Information System (INIS)

    Tsuji, Kouichi; Nakano, Kazuhiko; Ding Xunliang

    2007-01-01

    A new confocal micro X-ray fluorescence instrument was developed. This instrument has two independent micro X-ray tubes with Mo targets. A full polycapillary X-ray lens was attached to each X-ray tube. Another half polycapillary lens was attached to a silicon drift X-ray detector (SDD). The focal spots of the three lenses were adjusted to a common position. The effects of the excitation of two X-ray beams were investigated. The instrument enabled highly sensitive three-dimensional X-ray fluorescence analysis. We confirmed that the X-ray fluorescence intensity from the sample increased by applying the two independent X-ray tubes in confocal configuration. Elemental depth profiling of black wheat was demonstrated with the result that each element in the surface coat of a wheat grain showed unique distribution

  17. Collimating device for x-rays

    International Nuclear Information System (INIS)

    Schroeder, C.H.

    1978-01-01

    A position indicating device which attaches to an X-ray machine is provided with a collar onto which a lead lined cylinder may be threaded. The cylinder has an end plate through which a rectangular opening is formed in order to provide a rectangular X-ray beam for use in cephalometric radiography. When the cylinder is removed, a circular beam is emitted for use in intra-oral radiography

  18. UHV photoelectron x-ray beam position monitor

    International Nuclear Information System (INIS)

    Johnson, E.D.; Oversluizen, T.

    1989-01-01

    As part of our research program to develop viable beam position monitors for both the X-ray and VUV beamlines at the NSLS, we have constructed vertical photon beam position monitors which are presently mounted in two front-ends in the X-ray ring. These area-type detectors are located before the safety shutters and are, therefore, able to monitor the beam position even during injection. The features of this type of monitor which contribute to its long-term stability, position sensitivity, and immunity to horizontal beam motion have been examined and will be discussed. 6 refs., 4 figs

  19. Stabilization of synchrotron radiation x-ray beam by MOSTAB

    CERN Document Server

    Kudo, T P; Tanida, H; Furukawa, Y; Hirono, T; Ishikawa, T; Nishino, Y

    2003-01-01

    Monochromator stabilization (MOSTAB) is a feedback control system to stabilize an x-ray beam of synchrotron radiation. It applies a feedback voltage to a piezo electric transducer attached to a double-crystal monochromator. We developed MOSTAB modules and examined their performances using SPring-8 beamlines. The x-ray beam position stabilization using MOSTAB was realized simultaneously with the x-ray beam intensity stabilization. As an example of its application, we performed EXAFS measurement with MOSTAB. (author)

  20. Comparative analysis of partial imaging performance parameters of home and imported X-ray machines

    International Nuclear Information System (INIS)

    Cao Yunxi; Wang Xianyun; Liu Huiqin; Guo Yongxin

    2002-01-01

    Objective: To compare and analyze the performance indexes and the imaging quality of the home and imported X-ray machines through testing their partial imaging performance parameters. Methods: By separate sampling from 10 home and 10 imported X-ray machines, the parameters including tube current, time of exposure, machine total exposure, and repeatability were tested, and the imaging performance was evaluated according to the national standard. Results: All the performance indexes met the standard of GB4505-84. The first sampling tests showed the maximum changing coefficient of imaging performance repeatability of the home X-ray machines was Δmax1 = 0.025, while that of the imported X-ray machine was Δmax1 = 0.016. In the second sampling tests, the maximum changing coefficients of the two were Δmax2 = 0.048 and Δmax2 = 0.022, respectively. Conclusion: The 2 years' follow-up tests indicate that there is no significant difference between the above-mentioned parameters of the elaborately adjusted home X-ray machines and imported ones, but the home X-ray machines are no better than the imported X-ray machines in stability and consistency

  1. X-ray beam size measurements on the Advanced Test Accelerator

    International Nuclear Information System (INIS)

    Struve, K.W.; Chambers, F.W.; Lauer, E.J.; Slaughter, D.R.

    1986-01-01

    The electron beam size has been determined on the Advanced Test Accelerator (ATA) by intercepting the beam with a target and measuring the resulting x-ray intensity as a function of time as the target is moved through the beam. Several types of targets have been used. One is a tantalum rod which extends completely across the drift chamber. Another is a tungsten powder filled carbon crucible. Both of these probes are moved from shot to shot so that the x-ray signal intensity varies with probe position. A third is a larger tantalum disk which is inserted on beam axis to allow determining beam size on a one shot basis. The x-ray signals are detected with an MCP photomultiplier tube located at 90 0 to the beamline. It is sufficiently shielded to reject background x-rays and neutrons. The signals were digitized, recorded and later unfolded to produce plots of x-ray intensity versus probe position for several times during the pulse. The presumption that the x-ray intensity is proportional to beam current density is checked computationally. Details of the probe construction and PMT shielding, as well as sample measurements are given

  2. Design and performance of coded aperture optical elements for the CESR-TA x-ray beam size monitor

    Energy Technology Data Exchange (ETDEWEB)

    Alexander, J.P.; Chatterjee, A.; Conolly, C.; Edwards, E.; Ehrlichman, M.P. [Cornell University, Ithaca, NY 14853 (United States); Flanagan, J.W. [High Energy Accelerator Research Organization (KEK), Tsukuba (Japan); Department of Accelerator Science, Graduate University for Advanced Studies (SOKENDAI), Tsukuba (Japan); Fontes, E. [Cornell University, Ithaca, NY 14853 (United States); Heltsley, B.K., E-mail: bkh2@cornell.edu [Cornell University, Ithaca, NY 14853 (United States); Lyndaker, A.; Peterson, D.P.; Rider, N.T.; Rubin, D.L.; Seeley, R.; Shanks, J. [Cornell University, Ithaca, NY 14853 (United States)

    2014-12-11

    We describe the design and performance of optical elements for an x-ray beam size monitor (xBSM), a device measuring e{sup +} and e{sup −} beam sizes in the CESR-TA storage ring. The device can measure vertical beam sizes of 10–100μm on a turn-by-turn, bunch-by-bunch basis at e{sup ±} beam energies of ∼2–5GeV. x-rays produced by a hard-bend magnet pass through a single- or multiple-slit (coded aperture) optical element onto a detector. The coded aperture slit pattern and thickness of masking material forming that pattern can both be tuned for optimal resolving power. We describe several such optical elements and show how well predictions of simple models track measured performances. - Highlights: • We characterize optical element performance of an e{sup ±} x-ray beam size monitor. • We standardize beam size resolving power measurements to reference conditions. • Standardized resolving power measurements compare favorably to model predictions. • Key model features include simulation of photon-counting statistics and image fitting. • Results validate a coded aperture design optimized for the x-ray spectrum encountered.

  3. Innovation of High Voltage Supply Adjustment Device on Diagnostic X-Ray Machine

    International Nuclear Information System (INIS)

    Sujatno; Wiranto Budi Santoso

    2010-01-01

    Innovation of high voltage supply adjustment device on diagnostic x-ray machine has been carried out. The innovation is conducted by utilizing an electronic circuit as a high voltage adjustment device. Usually a diagnostic x-ray machine utilizes a transformer or an auto-transformer as a high voltage supply adjustment device. A high power diagnostic x-ray machine needs a high power transformer which has big physical dimension. Therefore a box control where the transformer is located has to have big physical dimension. Besides, the price of the transformer is expensive and hardly found in local markets. In this innovation, the transformer is replaced by an electronic circuit. The main component of the electronic circuit is Triac BTA-40. As adjustment device, the triac is controlled by a variable resistor which is coupled by a stepper motor. A step movement of stepper motor varies a value of resistor. The resistor value determines the triac gate voltage. Furthermore the triac will open according to the value of electrical current flowing to the gate. When the gate is open, electrical voltage and current will flow from cathode to anode of the triac. The value of these electrical voltage and current depend on gate open condition. Then this triac output voltage is feed to diagnostic x-ray machine high voltage supply. Therefore the high voltage value of diagnostic x-ray machine is adjusted by the output voltage of the electronic circuit. By using this electronic circuit, the physical dimension of diagnostic x-ray machine box control and the price of the equipment can be reduced. (author)

  4. Polychromatic X-ray Micro- and Nano-Beam Science and Instrumentation

    Science.gov (United States)

    Ice, G. E.; Larson, B. C.; Liu, W.; Barabash, R. I.; Specht, E. D.; Pang, J. W. L.; Budai, J. D.; Tischler, J. Z.; Khounsary, A.; Liu, C.; Macrander, A. T.; Assoufid, L.

    2007-01-01

    Polychromatic x-ray micro- and nano-beam diffraction is an emerging nondestructive tool for the study of local crystalline structure and defect distributions. Both long-standing fundamental materials science issues, and technologically important questions about specific materials systems can be uniquely addressed. Spatial resolution is determined by the beam size at the sample and by a knife-edge technique called differential aperture microscopy that decodes the origin of scattering from along the penetrating x-ray beam. First-generation instrumentation on station 34-ID-E at the Advanced Photon Source (APS) allows for nondestructive automated recovery of the three-dimensional (3D) local crystal phase and orientation. Also recovered are the local elastic-strain and the dislocation tensor distributions. New instrumentation now under development will further extend the applications of polychromatic microdiffraction and will revolutionize materials characterization.

  5. Polychromatic X-ray Micro- and Nano-Beam Science and Instrumentation

    International Nuclear Information System (INIS)

    Ice, G.E.; Larson, Ben C.; Liu, Wenjun; Barabash, Rozaliya; Specht, Eliot D; Pang, Judy; Budai, John D.; Tischler, Jonathan Zachary; Khounsary, Ali; Liu, Chian; Macrander, Albert T.; Assoufid, Lahsen

    2007-01-01

    Polychromatic x-ray micro- and nano-beam diffraction is an emerging nondestructive tool for the study of local crystalline structure and defect distributions. Both long-standing fundamental materials science issues, and technologically important questions about specific materials systems can be uniquely addressed. Spatial resolution is determined by the beam size at the sample and by a knife-edge technique called differential aperture microscopy that decodes the origin of scattering from along the penetrating x-ray beam. First-generation instrumentation on station 34-ID-E at the Advanced Photon Source (APS) allows for nondestructive automated recovery of the three-dimensional (3D) local crystal phase and orientation. Also recovered are the local elastic-strain and the dislocation tensor distributions. New instrumentation now under development will further extend the applications of polychromatic microdiffraction and will revolutionize materials characterization

  6. A microcontroller application as X-ray machine's high voltage controller

    International Nuclear Information System (INIS)

    Wiranto Budi Santoso; Beny Syawaludin

    2010-01-01

    A micro controller application as x-ray machine's high voltage controller has been carried out. The purpose of this micro controller application is to give an accurate high voltage supply to the x-ray tube so that the x-ray machine could produce the result as expected. The micro controller based X-ray machine's high voltage controller receives an input voltage from the keypad. This input value is displayed in the LCD (Liquid Crystal Display) screen. Then micro controller uses this input data to drive a stepper motor. The stepper motor adjusts the high voltage auto transformer's output according to the input value. The micro controller is programmed using BASCOM-8051 compiler. The test results show that the stepper motor could rotate according to an input value (author)

  7. A submicron synchrotron X-ray beam generated by capillary optics

    International Nuclear Information System (INIS)

    Engstroem, P.; Larsson, S.; Rindby, A.; Buttkewitz, A.; Garbe, S.; Gaul, G.; Knoechel, A.; Lechtenberg, F.; Deutsches Elektronen-Synchrotron

    1991-01-01

    A novel capillary optics technique for focusing synchrotron X-ray beams has been applied in an experiment performed at the DORIS storage ring at HASYLAB. This new technqiue, which utilizes the total reflection properties of X-rays inside small capillaries, has recently been applied to generate microbeams of X-rays, with a beam size down to about 10 μm using conventional X-ray tubes. The result from our recent experiment shows that capillary optics can also be used to generate a submicron beam of X-rays from a synchrotron light source. A description of the capillary unit, and the alignment procedure is given. The influence of the thermal load on the device caused by the intense flux of synchrotron radiation will be discussed. Future perspectives of the capillary techniques as applied to synchrotron radiation will be discussed. (orig.)

  8. A new beamstop for microfocus X-ray capillary beams

    Energy Technology Data Exchange (ETDEWEB)

    Englich, Ulrich, E-mail: ue22@cornell.edu [Cornell University, Macromolecular Diffraction at CHESS (MacCHESS), Cornell High Energy Synchrotron Source, 200L Wilson Laboratory, Ithaca, NY 14853 (United States); Revesz, Peter [Cornell University, Cornell High Energy Synchrotron Source, Ithaca, NY 14853 (United States); Miller, William [Cornell University, Macromolecular Diffraction at CHESS (MacCHESS), Cornell High Energy Synchrotron Source, 200L Wilson Laboratory, Ithaca, NY 14853 (United States)

    2011-09-01

    In order to accurately measure the photon flux and to assist in aligning the beam, we have designed a modified beamstop device based on a photo diode integrated with the beamstop. The beamstop contains a small CdWO{sub 4} crystal that completely stops the X-rays and at the same time produces photoluminescence proportional to the X-ray flux. The light is then guided to a photosensitive diode using a flexible light pipe to monitor the flux. With this device we achieve the goal of stopping the primary X-ray beam and simultaneously monitoring the X-ray intensity, thus eliminating the need for integrating ion-chambers into the capillary or collimator mount.

  9. Design of a New Collimation System to Prevent Interference between X-ray Machines and Radiation Portal Monitors

    International Nuclear Information System (INIS)

    Guzzardo, Tyler; Livesay, Jake

    2012-01-01

    Researchers at Oak Ridge National Laboratory (ORNL) developed a new collimation system that allows radiation portal monitors (RPMs) installed near x-ray machines to operate with a negligible false-positive alarm rate. RPMs are usually installed as far as possible from x-ray machines because false alarms are triggered by escaping x-rays; however, constraints at the installation site sometimes make it necessary that RPMs be installed near x-ray machines. Such RPMs are often plagued by high alarm rates resulting from the simultaneous operation of the RPMs and x-ray machines. Limitations on pedestrian flow, x-ray machine orientation, and RPM location often preclude a simple solution for lowering the alarm rate. Adding additional collimation to the x-ray machines to stop the x-rays at the source can reduce the alarm rate without interfering with site operations or adversely affecting the minimum detectable quantity of material (MDQ). A collimation design has been verified by measurements conducted at a RPM installation site and is applicable to all new and existing RPM installations near x-ray machines.

  10. Cone beam x-ray luminescence computed tomography: a feasibility study.

    Science.gov (United States)

    Chen, Dongmei; Zhu, Shouping; Yi, Huangjian; Zhang, Xianghan; Chen, Duofang; Liang, Jimin; Tian, Jie

    2013-03-01

    The appearance of x-ray luminescence computed tomography (XLCT) opens new possibilities to perform molecular imaging by x ray. In the previous XLCT system, the sample was irradiated by a sequence of narrow x-ray beams and the x-ray luminescence was measured by a highly sensitive charge coupled device (CCD) camera. This resulted in a relatively long sampling time and relatively low utilization of the x-ray beam. In this paper, a novel cone beam x-ray luminescence computed tomography strategy is proposed, which can fully utilize the x-ray dose and shorten the scanning time. The imaging model and reconstruction method are described. The validity of the imaging strategy has been studied in this paper. In the cone beam XLCT system, the cone beam x ray was adopted to illuminate the sample and a highly sensitive CCD camera was utilized to acquire luminescent photons emitted from the sample. Photons scattering in biological tissues makes it an ill-posed problem to reconstruct the 3D distribution of the x-ray luminescent sample in the cone beam XLCT. In order to overcome this issue, the authors used the diffusion approximation model to describe the photon propagation in tissues, and employed the sparse regularization method for reconstruction. An incomplete variables truncated conjugate gradient method and permissible region strategy were used for reconstruction. Meanwhile, traditional x-ray CT imaging could also be performed in this system. The x-ray attenuation effect has been considered in their imaging model, which is helpful in improving the reconstruction accuracy. First, simulation experiments with cylinder phantoms were carried out to illustrate the validity of the proposed compensated method. The experimental results showed that the location error of the compensated algorithm was smaller than that of the uncompensated method. The permissible region strategy was applied and reduced the reconstruction error to less than 2 mm. The robustness and stability were then

  11. Simple, compact, high brightness source for x-ray lithography and x-ray radiography

    International Nuclear Information System (INIS)

    Hawryluk, A.M.

    1986-01-01

    A simple, compact, high brightness x-ray source has recently been built. This source utilizes a commercially available, cylindrical geometry electron beam evaporator, which has been modified to enhance the thermal cooling to the anode. Cooling is accomplished by using standard, low-conductivity laboratory water, with an inlet pressure of less than 50 psi, and a flow rate of approx.0.3 gal/min. The anode is an inverted cone geometry for efficient cooling. The x-ray source has a measured sub-millimeter spot size (FWHM). The anode has been operated at 1 KW e-beam power (10 KV, 100 ma). Higher operating levels will be investigated. A variety of different x-ray lines can be obtained by the simple interchange of anodes of different materials. Typical anodes are made from easily machined metals, or materials which are vacuum deposited onto a copper anode. Typically, a few microns of material is sufficient to stop 10 KV electrons without significantly decreasing the thermal conductivity through the anode. The small size and high brightness of this source make it useful for step and repeat exposures over several square centimeter areas, especially in a research laboratory environment. For an aluminum anode, the estimated Al-K x-ray flux at 10 cms from the source is 70 μW/cm 2

  12. Modification of a Superficial X-Ray Therapy Machine for Rectal Contact Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Barish, Robert J.; Donohue, Karen Episcopia

    2015-01-15

    X-ray therapy of superficial rectal cancers using a hand-held 50 kV contact unit (Philips RT-50) in a technique first described by Papillon had reached a point of widening clinical acceptability when the manufacturer of this equipment discontinued its production. To pursue this endocavitary approach to rectal therapy, technical modifications have to be made to conventional superficial x-ray therapy machines. Advantages over the original Papillon method include remote viewing of the lesion through the proctoscopic cone and a lower radiation exposure for the operator. We have evaluated a Bucky Combination Therapy Unit under conditions in which the operating voltage (65 kV), target skin distance (23.6 cm), and added filtration (0.39 mm Al) were selected in order to match as closely as possible the beam penetration characteristics of the “standard” (Papillon) technique. With this equipment, the thermal characteristics of the tube anode and housing limit the amount of radiation that can be delivered before a “rest period” for the machine is needed. In practice, 3 minutes of irradiation at an exposure rate of 500 R/min can be performed followed by an interval of 3 minutes before irradiation can be resumed.

  13. Constancy check of beam quality in conventional diagnostic X-ray equipment

    International Nuclear Information System (INIS)

    Costa, Alessandro M.; Badin, Romulo S.; Leite, Marina S.; Caldas, Linda V.E.

    2008-01-01

    A tandem ionization chamber was developed for quality control programs of X-ray equipment used in conventional radiography and mammography. A methodology for the use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities was established. The application at a medical X-ray imaging facility of this established methodology is presented. The use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities is a useful method to control the performance of the X-ray equipment

  14. Establishing the standard X-ray beam qualities for calibration of dosimeters used in diagnostic radiology following IAEA-TRS457

    International Nuclear Information System (INIS)

    Duong Van Trieu; Ho Quang Tuan; Bui Duc Ky

    2014-01-01

    The determination of the patient dose needs to provide a reference dose for the patient that reference dose levels to assess the relative risk during X- ray diagnostic. This mission, We had established a number of standard beam qualities to perform calibrations of diagnostic dosimeters and methods of measuring patient dose in X-ray diagnostic. At radiation dosimetry room, we had establish RQR2, RQR3, RQR4, RQR5, RQR6 beam qualities based on IAEA-TRS457 documentation with homogeneity coefficient (h) for each beam quality in the range 0.7 - 0.8, and haft-value layers HVL1, HVL2 of experimental and IAEA is different about 10%. Established calibration method for diagnostic dosimeters as KAP meters, UNFORS dosimeters, and the TLD dosimeters, practical measurements of entrance surface air kerma on Shimadzu X-ray machines used phantom. (author)

  15. Single-shot beam-position monitor for x-ray free electron laser

    Science.gov (United States)

    Tono, Kensuke; Kudo, Togo; Yabashi, Makina; Tachibana, Takeshi; Feng, Yiping; Fritz, David; Hastings, Jerome; Ishikawa, Tetsuya

    2011-02-01

    We have developed an x-ray beam-position monitor for detecting the radiation properties of an x-ray free electron laser (FEL). It is composed of four PIN photodiodes that detect backscattered x-rays from a semitransparent diamond film placed in the beam path. The signal intensities from the photodiodes are used to compute the beam intensity and position. A proof-of-principle experiment at a synchrotron light source revealed that the error in the beam position is reduced to below 7 μm by using a nanocrystal diamond film prepared by plasma-enhanced chemical vapor deposition. Owing to high dose tolerance and transparency of the diamond film, the monitor is suitable for routine diagnostics of extremely intense x-ray pulses from the FEL.

  16. The development of a postal method to assess X-ray beam parameters and image quality in dental radiology

    International Nuclear Information System (INIS)

    Fenton, D.M.

    1994-10-01

    Intraoral radiographs are an extremely valuable diagnostic tool in dentistry. Radiography permits the early detection and diagnosis of dental disease and consequently is used extensively. However, public concern about radiation exposure has increased in recent times. This concern is reflected in national and international law, to the extent that, the basic principles of radiological protection, that is, justification, optimisation and dose limitation are written into law. Furthermore, in Ireland, the regulations, as outlined in the Code of Practice for Radiological Protection in Dentistry, require intraoral dental X-ray machines to perform to certain standards. A report of a direct survey of 164 intraoral dental X-ray machines is given in this study. The survey covered mechanical, electrical as well as radiation safety. Inadequacies with respect to focus to skin distance and timer accuracy were found in 45% and 42% of the machines surveyed. Ninety eight machines were assessed for electrical safety in which 48% were found to be unsafe. The results indicate that a complete assessment of the performance of dental X-ray units in Ireland is required. However, as there are in excess of 800 dental X-ray machines located throughout the country, such an assessment would be very costly for the regulatory authority. The development of a postal method for the assessment of the performance of dental X-ray machines is described in this study. This postal method provides information on the kV, total filtration, beam width and timer linearity and is undertaken by means of a penetrameter and film envelopes for exposure to the X-ray set under examination, together with a questionnaire that requests information on environment in which the machine is located. Using this method an accuracy of +-5% of the actual value was achieved in the measurement of kVp. The penetrameter was also used to assess whether or not the filtration of a particular machine complies with the regulations. This

  17. Hard X-ray bremsstrahlung production in solar flares by high-energy proton beams

    Science.gov (United States)

    Emslie, A. G.; Brown, J. C.

    1985-01-01

    The possibility that solar hard X-ray bremsstrahlung is produced by acceleration of stationary electrons by fast-moving protons, rather than vice versa, as commonly assumed, was investigated. It was found that a beam of protons which involves 1836 times fewer particles, each having an energy 1836 times greater than that of the electrons in the equivalent electron beam model, has exactly the same bremsstrahlung yield for a given target, i.e., the mechanism has an energetic efficiency equal to that of conventional bremsstrahlung models. Allowance for the different degrees of target ionization appropriate to the two models (for conventional flare geometries) makes the proton beam model more efficient than the electron beam model, by a factor of order three. The model places less stringent constraints than a conventional electron beam model on the flare energy release mechanism. It is also consistent with observed X-ray burst spectra, intensities, and directivities. The altitude distribution of hard X-rays predicted by the model agrees with observations only if nonvertical injection of the protons is assumed. The model is inconsistent with gamma-ray data in terms of conventional modeling.

  18. A novel process control method for a TT-300 E-Beam/X-Ray system

    Science.gov (United States)

    Mittendorfer, Josef; Gallnböck-Wagner, Bernhard

    2018-02-01

    This paper presents some aspects of the process control method for a TT-300 E-Beam/X-Ray system at Mediscan, Austria. The novelty of the approach is the seamless integration of routine monitoring dosimetry with process data. This allows to calculate a parametric dose for each production unit and consequently a fine grain and holistic process performance monitoring. Process performance is documented in process control charts for the analysis of individual runs as well as historic trending of runs of specific process categories over a specified time range.

  19. The reduction methods of operator's radiation dose for portable dental X-ray machines.

    Science.gov (United States)

    Cho, Jeong-Yeon; Han, Won-Jeong

    2012-08-01

    This study was aimed to investigate the methods to reduce operator's radiation dose when taking intraoral radiographs with portable dental X-ray machines. Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used. Operator's radiation dose was measured with an 1,800 cc ionization chamber (RadCal Corp.) at the hand level of X-ray tubehead and at the operator's chest and waist levels with and without the backscatter shield. The operator's radiation dose at the hand level was measured with and without lead gloves and with long and short cones. The backscatter shield reduced operator's radiation dose at the hand level of X-ray tubehead to 23 - 32%, the lead gloves to 26 - 31%, and long cone to 48 - 52%. And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%. When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

  20. Estimating The Reliability of the Lawrence Livermore National Laboratory (LLNL) Flash X-ray (FXR) Machine

    International Nuclear Information System (INIS)

    Ong, M M; Kihara, R; Zentler, J M; Kreitzer, B R; DeHope, W J

    2007-01-01

    At Lawrence Livermore National Laboratory (LLNL), our flash X-ray accelerator (FXR) is used on multi-million dollar hydrodynamic experiments. Because of the importance of the radiographs, FXR must be ultra-reliable. Flash linear accelerators that can generate a 3 kA beam at 18 MeV are very complex. They have thousands, if not millions, of critical components that could prevent the machine from performing correctly. For the last five years, we have quantified and are tracking component failures. From this data, we have determined that the reliability of the high-voltage gas-switches that initiate the pulses, which drive the accelerator cells, dominates the statistics. The failure mode is a single-switch pre-fire that reduces the energy of the beam and degrades the X-ray spot-size. The unfortunate result is a lower resolution radiograph. FXR is a production machine that allows only a modest number of pulses for testing. Therefore, reliability switch testing that requires thousands of shots is performed on our test stand. Study of representative switches has produced pre-fire statistical information and probability distribution curves. This information is applied to FXR to develop test procedures and determine individual switch reliability using a minimal number of accelerator pulses

  1. Beam line for experiments with coherent soft x-rays

    International Nuclear Information System (INIS)

    Howells, M.R.; Kirz, J.; Krinsky, S.

    1982-12-01

    The advantages of coherent soft x-rays for three-dimensional imaging of biological specimens are discussed, the x-ray source requirements are described, and the general design of the beam line and its optical system are given

  2. X-ray topography using the forward transmitted beam under multiple-beam diffraction conditions

    Energy Technology Data Exchange (ETDEWEB)

    Tsusaka, Y., E-mail: tsusaka@sci.u-hyogo.ac.jp; Takano, H. [Graduate School of Material Science, University of Hyogo, 3-2-1, Kouto, Kamigori, Hyogo 678-1297 (Japan); Takeda, S. [SPring-8 Service Co., Ltd., 1-20-5, Kouto, Shingu, Tatsuno, Hyogo 679-5165 (Japan); Yokoyama, K.; Matsui, J. [Synchrotron Radiation Nanotechnology Center, University of Hyogo, 1-490-2, Kouto, Shingu, Tatsuno, Hyogo 679-5165 (Japan); Kagoshima, Y. [Graduate School of Material Science, University of Hyogo, 3-2-1, Kouto, Kamigori, Hyogo 678-1297 (Japan); Synchrotron Radiation Nanotechnology Center, University of Hyogo, 1-490-2, Kouto, Shingu, Tatsuno, Hyogo 679-5165 (Japan)

    2016-02-15

    X-ray topographs are taken for a sapphire wafer with the [0001] surface normal, as an example, by forward transmitted synchrotron x-ray beams combined with two-dimensional electronic arrays in the x-ray detector having a spatial resolution of 1 μm. They exhibit no shape deformation and no position shift of the dislocation lines on the topographs. Since the topography is performed under multiple-beam diffraction conditions, the topographic images of a single diffraction (two-wave approximation condition) or plural diffractions (six-wave approximation condition) can be recorded without large specimen position changes. As usual Lang topographs, it is possible to determine the Burgers vector of each dislocation line. Because of high parallelism of the incoming x-rays and linear sensitivity of the electronic arrays to the incident x-rays, the present technique can be used to visualize individual dislocations in single crystals of the dislocation density as high as 1 × 10{sup 5} cm{sup −2}.

  3. An x-ray technique for precision laser beam synchronization

    International Nuclear Information System (INIS)

    Landen, O.L.; Lerche, R.A.; Hay, R.G.; Hammel, B.A.; Kalantar, D.; Cable, M.D.

    1994-01-01

    A new x-ray technique for recording the relative arrival times of multiple laser beams at a common target with better than ± 10 ps accuracy has been implemented at the Nova laser facility. 100 ps, 3ω Nova beam are focused to separate locations on a gold ribbon target viewed from the side. The measurement consists of using well characterized re-entrant x-ray streak cameras for 1-dimensional streaked imaging of the > 3 keV x-rays emanating from these isolated laser plasmas. After making the necessary correction for the differential laser, x-ray and electron transit times involved, timing offsets as low as ± 7 ps are resolved, and on subsequent shots, corrected for, verified and independently checked. This level of synchronization proved critical in meeting the power balance requirements for indirectly-driven pulse-shaped Nova implosions

  4. A Study on Mono-energetic Beam Source Using Characteristic X-ray for Substance Identification System

    International Nuclear Information System (INIS)

    Lee, Hwan Soo

    2009-02-01

    A new mono-energetic beam source was developed by using characteristic X-ray for improving performance of the substance identification system. Most of inspection systems use X-ray tubes for their source modules. However, the broad energy spectrum of X-ray tube causes an increase of uncertainty. In this study, it was found that mono-energetic beam sources can be generated by using X-ray tube and the designed target filter assembly. In order to investigate the monoenergetic beam source, the sensitivity study was conducted with a series of different X-ray tube potentials, radiator and filter materials using Monte Carlo simulation. The developed beam sources have a mono-energy peak at 69 keV, 78 keV and 99 keV, and they are named as characteristic X-ray beam BEAM69, BEAM78 and BEAM99, respectively. The characteristic X-ray beam intensity was over thirty three times more than that of hardening beam used previous work at Hanyang University. And BEAM69 and BEAM99 were applied to the substance identification system as a source. The relative error between results of characteristic X-ray beams and 69 keV and 99 keV photons was about 2% on the average for five unknown materials. In comparison with experiment results by using hardening beam, characteristic X-ray beam achieves better accuracy which is about 6.46 % on the average. Hence, it is expected that the developed characteristic X-ray beam source helps lower uncertainty of the inspection system, and the inspection time will be reduced considerably due to its high beam intensity

  5. Design of parallel dual-energy X-ray beam and its performance for security radiography

    International Nuclear Information System (INIS)

    Kim, Kwang Hyun; Myoung, Sung Min; Chung, Yong Hyun

    2011-01-01

    A new concept of dual-energy X-ray beam generation and acquisition of dual-energy security radiography is proposed. Erbium (Er) and rhodium (Rh) with a copper filter were positioned in front of X-ray tube to generate low- and high-energy X-ray spectra. Low- and high-energy X-rays were guided to separately enter into two parallel detectors. Monte Carlo code of MCNPX was used to derive an optimum thickness of each filter for improved dual X-ray image quality. It was desired to provide separation ability between organic and inorganic matters for the condition of 140 kVp/0.8 mA as used in the security application. Acquired dual-energy X-ray beams were evaluated by the dual-energy Z-map yielding enhanced performance compared with a commercial dual-energy detector. A collimator for the parallel dual-energy X-ray beam was designed to minimize X-ray beam interference between low- and high-energy parallel beams for 500 mm source-to-detector distance.

  6. Dosimetry of x-ray beams: The measure of the problem

    International Nuclear Information System (INIS)

    de Castro, T.M.

    1986-08-01

    This document contains the text of an oral presentation on dosimetry of analytical x-ray equipment presented at the Denver X-Ray Conference. Included are discussions of sources of background radiation, exposure limits from occupational sources, and the relationship of these sources to the high dose source of x-rays found in analytical machines. The mathematical basis of x-ray dosimetry is reviewed in preparation for more detailed notes on personnel dosimetry and the selection of the most appropriate dosimeter for a specific application. The presentation concludes with a discussion common to previous x-ray equipment accidents. 2 refs

  7. Generation of linearly polarized resonant transition radiation X-ray beam

    International Nuclear Information System (INIS)

    Yajima, Kazuaki; Awata, Takaaki; Ikeda, Mitsuharu; Ikeda, Kenichi; Yogo, Akifumi; Itoh, Akio; Imanishi, Nobutsugu

    2000-01-01

    We have proposed a method to generate almost linearly polarized resonant transition radiation X rays by using a rectangular slit placed on an electron beam axis. Our calculation predicted that the linearity is 93.5% for the resonant transition radiation X-ray beam extracted through a slit of 0.5 mrad long and 0.2 mrad wide in case of 1-GeV electron beam irradiating a 7.5-μm thick Kapton foil stack. (author)

  8. Generation of linearly polarized resonant transition radiation X-ray beam

    Energy Technology Data Exchange (ETDEWEB)

    Yajima, Kazuaki; Awata, Takaaki; Ikeda, Mitsuharu; Ikeda, Kenichi; Yogo, Akifumi; Itoh, Akio; Imanishi, Nobutsugu [Kyoto Univ. (Japan). Dept. of Nuclear Engineering

    2000-03-01

    We have proposed a method to generate almost linearly polarized resonant transition radiation X rays by using a rectangular slit placed on an electron beam axis. Our calculation predicted that the linearity is 93.5% for the resonant transition radiation X-ray beam extracted through a slit of 0.5 mrad long and 0.2 mrad wide in case of 1-GeV electron beam irradiating a 7.5-{mu}m thick Kapton foil stack. (author)

  9. Comparation of Some Values of Measured Parameteres between Old and New X-Ray Machines

    International Nuclear Information System (INIS)

    Kocic, B.; Marinkovic, J.; Praskalo, J.

    2013-01-01

    Checking the change of values of parameters like specific value of kerma, repeatability of expositions, linearity of exposition, stability of high voltage, time of exposition and HVL of X-ray machines that are in use in medical centres of Republic of Serbia give us an idea of comparing those values for two groups of X-ray machines. One is in use since 70's in last century and is still in use, and the other group is quite new, and is controlled for last 5 years. Checking the results of controlled parameters within those two groups of X-ray machines we can compare stability and quality of old and new machines, and the influence of 'aging' of machines to the quality and stability of measured parameters. Some of conclusions are that old machines are showing better stability and repeatability of values of measured parameters during the 'aging' than new machines.(author)

  10. Wavelength dispersive X-ray absorption fine structure imaging by parametric X-ray radiation

    International Nuclear Information System (INIS)

    Inagaki, Manabu; Sakai, Takeshi; Sato, Isamu; Hayakawa, Yasushi; Nogami, Kyoko; Tanaka, Toshinari; Hayakawa, Ken; Nakao, Keisuke

    2008-01-01

    The parametric X-ray radiation (PXR) generator system at Laboratory for Electron Beam Research and Application (LEBRA) in Nihon University is a monochromatic and coherent X-ray source with horizontal wavelength dispersion. The energy definition of the X-rays, which depends on the horizontal size of the incident electron beam on the generator target crystal, has been investigated experimentally by measuring the X-ray absorption near edge structure (XANES) spectra on Cu and CuO associated with conventional X-ray absorption imaging technique. The result demonstrated the controllability of the spectrum resolution of XANES by adjusting of the horizontal electron beam size on the target crystal. The XANES spectra were obtained with energy resolution of several eV at the narrowest case, which is in qualitative agreement with the energy definition of the PXR X-rays evaluated from geometrical consideration. The result also suggested that the wavelength dispersive X-ray absorption fine structure measurement associated with imaging technique is one of the promising applications of PXR. (author)

  11. X-ray radiotherapy

    International Nuclear Information System (INIS)

    Tronc, D.

    1995-01-01

    Full text: The most common form of radio therapy is X-ray therapy, where a beam of photons or their parent electrons break down hydrogen bonds within the body's cells and remove certain DNA information necessary for cell multiplication. This process can eradicate malignant cells leading to complete recovery, to the remission of some cancers, or at least to a degree of pain relief. The radiotherapy instrument is usually an electron linac, and the electrons are used either directly in 'electrotherapy' for some 10% of patients, or the electrons bombard a conversion target creating a broad beam of high energy photons or 'penetration X-rays'. The simplest machine consists of several accelerating sections at around 3 GHz, accelerating electrons to 6 MeV; a cooled tungsten target is used to produce a 4 Gray/min X-ray field which can be collimated into a rectangular shape at the patient position. This tiny linac is mounted inside a rotating isocentric gantry above the patient who must remain perfectly still. Several convergent beams can also be used to increase the delivered dose. More sophisticated accelerators operate at up to 18 MeV to increase penetration depths and decrease skin exposure. Alternatively, electrotherapy can be used with different energies for lower and variable penetration depths - approximately 0.5 cm per MeV. In this way surface tissue may be treated without affecting deeper and more critical anatomical regions. This type of linac, 1 to 2 metres long, is mounted parallel to the patient with a bending magnet to direct the beam to the radiotherapy system, which includes the target, thick movable collimator jaws, a beam field equalizer, dose rate and optical field simulation and energy controls. There are over 2000 acceleratorbased X-ray treatment units worldwide. Western countries have up to two units per million population, whereas in developing countries such as Bangladesh, the density is only one per 100 million. Several

  12. Development of a multi-lane X-ray mirror providing variable beam sizes

    Energy Technology Data Exchange (ETDEWEB)

    Laundy, D., E-mail: david.laundy@diamond.ac.uk; Sawhney, K.; Nistea, I.; Alcock, S. G.; Pape, I.; Sutter, J.; Alianelli, L.; Evans, G. [Diamond Light Source Ltd., Harwell Science and Innovation Campus, Didcot OX11 0DE (United Kingdom)

    2016-05-15

    Grazing incidence mirrors are used on most X-ray synchrotron beamlines to focus, collimate or suppress harmonics. Increasingly beamline users are demanding variable beam shapes and sizes at the sample position. We have now developed a new concept to rapidly vary the beam size and shape of a focused X-ray beam. The surface of an elliptically figured mirror is divided into a number of laterally separated lanes, each of which is given an additional longitudinal height profile calculated to shape the X-ray beam to a top-hat profile in the focal plane. We have now fabricated two prototype mirrors and present the results of metrology tests and measurements made with one of the mirrors focusing the X-rays on a synchrotron beamline. We envisage that such mirrors could be widely applied to rapid beam-size switching on many synchrotron beamlines.

  13. Quality assurance for diagnostic X-ray machines in Tanzania

    International Nuclear Information System (INIS)

    Sungita, Y.Y.; Mdoe, S.L.; Ngatunga, J.; Kitosi, A.E.; Muhogora, W.E.

    1998-01-01

    In this presentation it is being discussed a close relationship between the prevention of accidents involving apparatus that generates the ionizing radiation (x-ray tubes etc.) and the need to perform quality control procedures and make follow-up corrective maintenance procedures. A summary of results of quality control performance of x-ray machines in certain centers in Tanzania is tabled and measures to overcome some of the observed problems are recommended. The number of X-ray units inspected were 219, out of which 123 were working, 75 were out of order, 6 units were not yet installed and those which are working with faulty parts were 15. The performance of quality assurance for X-Ray units (57) tested showed that 36.8% didn't qualify. About 80% of these equipment are more than 15 years old. (author)

  14. Is low-energy-ion bombardment generated X-ray emission a secondary mutational source to ion-beam-induced genetic mutation?

    Energy Technology Data Exchange (ETDEWEB)

    Thongkumkoon, P. [Plasma and Beam Physics Research Facility, Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200 (Thailand); Thailand Center of Excellence in Physics, Commission on Higher Education, 328 Si Ayutthaya Road, Bangkok 10400 (Thailand); Prakrajang, K. [Plasma and Beam Physics Research Facility, Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200 (Thailand); Faculty of Science, Maejo University, Chiang Mai 50290 (Thailand); Thopan, P.; Yaopromsiri, C. [Plasma and Beam Physics Research Facility, Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200 (Thailand); Suwannakachorn, D. [Plasma and Beam Physics Research Facility, Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200 (Thailand); Thailand Center of Excellence in Physics, Commission on Higher Education, 328 Si Ayutthaya Road, Bangkok 10400 (Thailand); Yu, L.D., E-mail: yuld@fnrf.science.cmu.ac.th [Plasma and Beam Physics Research Facility, Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200 (Thailand); Thailand Center of Excellence in Physics, Commission on Higher Education, 328 Si Ayutthaya Road, Bangkok 10400 (Thailand)

    2013-07-15

    Highlights: ► Detected X-ray emission from metal, plastic and biological samples. ► Characteristic X-ray emission was detected from metal but not from non-metals. ► Low-energy ion bombarded bacteria held in different sample holders. ► Bacteria held in metal holder had higher mutation rate than in plastic holder. ► Ion-beam-induced X-ray from biological sample is not a basic mutation source. -- Abstract: Low-energy ion beam biotechnology has achieved tremendous successes in inducing crop mutation and gene transfer. However, mechanisms involved in the related processes are not yet well understood. In ion-beam-induced mutation, ion-bombardment-produced X-ray has been proposed to be one of the secondary mutation sources, but the speculation has not yet been experimentally tested. We carried out this investigation to test whether the low-energy ion-beam-produced X-ray was a source of ion-beam-induced mutation. In the investigation, X-ray emission from 29-keV nitrogen- or argon- ion beam bombarded bacterial Escherichia coli (E. coli) cells held in a metal or plastic sample holder was in situ detected using a highly sensitive X-ray detector. The ion beam bombarded bacterial cells held in different material holders were observed for mutation induction. The results led to a conclusion that secondary X-ray emitted from ion-beam-bombarded biological living materials themselves was not a, or at least a negligible, mutational source, but the ion-beam-induced X-ray emission from the metal that made the sample holder could be a source of mutation.

  15. Is low-energy-ion bombardment generated X-ray emission a secondary mutational source to ion-beam-induced genetic mutation?

    International Nuclear Information System (INIS)

    Thongkumkoon, P.; Prakrajang, K.; Thopan, P.; Yaopromsiri, C.; Suwannakachorn, D.; Yu, L.D.

    2013-01-01

    Highlights: ► Detected X-ray emission from metal, plastic and biological samples. ► Characteristic X-ray emission was detected from metal but not from non-metals. ► Low-energy ion bombarded bacteria held in different sample holders. ► Bacteria held in metal holder had higher mutation rate than in plastic holder. ► Ion-beam-induced X-ray from biological sample is not a basic mutation source. -- Abstract: Low-energy ion beam biotechnology has achieved tremendous successes in inducing crop mutation and gene transfer. However, mechanisms involved in the related processes are not yet well understood. In ion-beam-induced mutation, ion-bombardment-produced X-ray has been proposed to be one of the secondary mutation sources, but the speculation has not yet been experimentally tested. We carried out this investigation to test whether the low-energy ion-beam-produced X-ray was a source of ion-beam-induced mutation. In the investigation, X-ray emission from 29-keV nitrogen- or argon- ion beam bombarded bacterial Escherichia coli (E. coli) cells held in a metal or plastic sample holder was in situ detected using a highly sensitive X-ray detector. The ion beam bombarded bacterial cells held in different material holders were observed for mutation induction. The results led to a conclusion that secondary X-ray emitted from ion-beam-bombarded biological living materials themselves was not a, or at least a negligible, mutational source, but the ion-beam-induced X-ray emission from the metal that made the sample holder could be a source of mutation

  16. Application Research of Quasi-monochromatic X-ray Machine Replacing 241Am Radioactive Source in Thickness Measurement

    Directory of Open Access Journals (Sweden)

    LIN Hui1;XIAO Xue-fu2;HOU Yue-xin3;ZHAO Jing1;JIAN Li-min1

    2014-02-01

    Full Text Available In the research of BS-03 thickness detector, a self-designed quasi-monochromatic (50-60 keV X-ray machine was studied as a substitution to Am-241 low energy photon source(1.11×109 Bq)The range of output current, the attenuation of X rays through the steel plate, the aluminum plate, and the organic glass plate and stability of quasi-monochromatic X-ray machine was tested. The result showed that quasi-monochromatic X-ray machine could be an applicable replacement of Am-241 radioactive source.

  17. Anomalous x-ray radiation of beam plasma

    International Nuclear Information System (INIS)

    Dimitrov, S.K.; Zavyalov, M.A.; Mikhin, S.G.; Tarasenkov, V.A.; Telkovskij, V.G.; Khrabrov, V.A.

    1985-01-01

    The properties of non-equilibrium stationary plasma under the conditions of the planned plasma-chemical reactors based on beam-plasma discharge were investigated. The x-ray spectrum of the beam-plasma was measured and anomalous spectral properties were analyzed. Starting with some critical pressure the anomalous radiation was added to the classical bremsstrahlung spectrum. The occurrence of anomalous radiation can be used to diagnose the condition of beam transportation in such systems. (D.Gy.)

  18. X-ray machine vision and computed tomography

    International Nuclear Information System (INIS)

    Anon.

    1988-01-01

    This survey examines how 2-D x-ray machine vision and 3-D computed tomography will be used in industry in the 1988-1995 timeframe. Specific applications are described and rank-ordered in importance. The types of companies selling and using 2-D and 3-D systems are profiled, and markets are forecast for 1988 to 1995. It is known that many machine vision and automation companies are now considering entering this field. This report looks at the potential pitfalls and whether recent market problems similar to those recently experienced by the machine vision industry will likely occur in this field. FTS will publish approximately 100 other surveys in 1988 on emerging technology in the fields of AI, manufacturing, computers, sensors, photonics, energy, bioengineering, and materials

  19. Cone-beam x-ray luminescence computed tomography based on x-ray absorption dosage

    Science.gov (United States)

    Liu, Tianshuai; Rong, Junyan; Gao, Peng; Zhang, Wenli; Liu, Wenlei; Zhang, Yuanke; Lu, Hongbing

    2018-02-01

    With the advances of x-ray excitable nanophosphors, x-ray luminescence computed tomography (XLCT) has become a promising hybrid imaging modality. In particular, a cone-beam XLCT (CB-XLCT) system has demonstrated its potential in in vivo imaging with the advantage of fast imaging speed over other XLCT systems. Currently, the imaging models of most XLCT systems assume that nanophosphors emit light based on the intensity distribution of x-ray within the object, not completely reflecting the nature of the x-ray excitation process. To improve the imaging quality of CB-XLCT, an imaging model that adopts an excitation model of nanophosphors based on x-ray absorption dosage is proposed in this study. To solve the ill-posed inverse problem, a reconstruction algorithm that combines the adaptive Tikhonov regularization method with the imaging model is implemented for CB-XLCT reconstruction. Numerical simulations and phantom experiments indicate that compared with the traditional forward model based on x-ray intensity, the proposed dose-based model could improve the image quality of CB-XLCT significantly in terms of target shape, localization accuracy, and image contrast. In addition, the proposed model behaves better in distinguishing closer targets, demonstrating its advantage in improving spatial resolution.

  20. X-ray diffraction and imaging with a coherent beam: application to X-ray optical elements and to crystals exhibiting phase inhomogeneities

    International Nuclear Information System (INIS)

    Masiello, F.

    2011-05-01

    The exceptional properties of synchrotron light sources have been exploited in very different disciplines, from archaeology to chemistry, from material science to biology, from medicine to physics. Among these properties it is important to mention the high brilliance, continuum spectrum, high degree of polarization, time structure, small source size and divergence of the beam, the last resulting in a high transversal coherence of the produced radiation. This high transversal coherence of the synchrotron sources has permitted the development of new techniques, e.g. phase contrast imaging, X-ray photon correlation spectroscopy and coherent X-ray diffraction imaging (CXDI). This thesis work will consist essentially of three parts. In the first part it will be presented the work done as a member of the X-ray Optics Group of ESRF in the characterization of high quality diamond crystals foreseen as X-ray optical elements. The characterization has been done using different complementary X-ray techniques, such as high resolution diffraction, topography, grazing incidence diffraction, reflectivity and measurements of the coherence preservation using the Talbot effect. In the second part, I will show the result obtained in the study of the temperature behaviours of the domain in periodically poled ferroelectrics crystals. This type of measurements, based on Bragg-Fresnel diffraction, are possible only thanks to the high degree of coherence of the beam. In the third part, I will present the results obtained in the characterization of diamonds foreseen for applications other than X-ray optical elements. (author)

  1. SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration

    Energy Technology Data Exchange (ETDEWEB)

    Wunderle, K [Cleveland Clinic Foundation, Cleveland, OH (United States); Wayne State University School of Medicine, Detroit, MI (United States); Godley, A; Shen, Z; Dong, F [Cleveland Clinic Foundation, Cleveland, OH (United States); Rakowski, J [Wayne State University School of Medicine, Detroit, MI (United States)

    2016-06-15

    Purpose: The purpose of this investigation is to characterize and quantify X-ray beam profiles for fluoroscopic x-ray beam spectra incorporating spectral (copper) filtration. Methods: A PTW (Freiburg, Germany) type 60016 silicon diode detector and PTW MP3 water tank were used to measure X-ray beam profiles for 60, 80, 100 and 120 kVp x-ray beams at five different copper filtration thicknesses ranging from 0–0.9 mm at 22 and 42 cm fields of view and depths of 1, 5, and 10 cm in both the anode-cathode axis (inplane) and cross-plane directions. All measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope inverted from the typical orientation providing an x-ray beam originating from above the water surface with the water level set at 60 cm from the focal spot. Results: X-ray beam profiles for beam spectra without copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. Our data collection benefited from the geometric orientation of the fluoroscope, providing a beam perpendicular to the tank water surface, rather than through a thin side wall as did the previously mentioned study. Profiles for beams with copper filtration were obtained which have not been previously investigated and published. Beam profiles in the anode-cathode axis near the surface and at lower x-ray energy exhibited substantial heel effect, which became less pronounced at greater depth. At higher energy with copper filtration in the beam, the dose falloff out-of-field became less pronounced, as would be anticipated given higher scatter photon energy. Conclusion: The x-ray beam profile data for the fluoroscopic x-ray beams incorporating copper filtration are intended for use as reference data for estimating doses to organs or soft tissue, including fetal dose, involving similar beam qualities or for comparison with mathematical models.

  2. Controlling X-ray beam trajectory with a flexible hollow glass fibre

    Energy Technology Data Exchange (ETDEWEB)

    Tanaka, Yoshihito, E-mail: yotanaka@riken.jp [RIKEN, 1-1-1 Kouto, Sayo-cho, Hyogo 679-5148 (Japan); Kwansei Gakuin University, Gakuen, Sanda, Hyogo 669-1337 (Japan); Nakatani, Takashi; Onitsuka, Rena [Kwansei Gakuin University, Gakuen, Sanda, Hyogo 669-1337 (Japan); Sawada, Kei [RIKEN, 1-1-1 Kouto, Sayo-cho, Hyogo 679-5148 (Japan); Takahashi, Isao [Kwansei Gakuin University, Gakuen, Sanda, Hyogo 669-1337 (Japan)

    2014-01-01

    X-ray beam trajectory control has been performed by using a 1.5 m-long flexible hollow glass fibre. A two-dimensional scan of a synchrotron radiation beam was demonstrated for X-ray absorption mapping. A metre-length flexible hollow glass fibre with 20 µm-bore and 1.5 mm-cladding diameters for transporting a synchrotron X-ray beam and controlling the trajectory has been examined. The large cladding diameter maintains a moderate curvature to satisfy the shallow glancing angle of total reflection. The observed transmission efficiency was more than 20% at 12.4 keV. As a demonstration, a wide-area scan of a synchrotron radiation beam was performed to identify the elements for a fixed metal film through its absorption spectra.

  3. Diagnostic X-ray spectrometry using a commercial CdZnTe detector

    International Nuclear Information System (INIS)

    Becker, P.H.B.

    1998-01-01

    X ray spectrometry using Ge or Si detectors is an established tool to measure characterization parameters of X-ray beams. This work describes how a commercial CdZnTe was used to perform diagnostic X-ray spectrometry. Spectra were measured for two X-ray machines and compared with similar data found in the literature with an agreement of 2% rms

  4. An MCNP-based model of a medical linear accelerator x-ray photon beam.

    Science.gov (United States)

    Ajaj, F A; Ghassal, N M

    2003-09-01

    The major components in the x-ray photon beam path of the treatment head of the VARIAN Clinac 2300 EX medical linear accelerator were modeled and simulated using the Monte Carlo N-Particle radiation transport computer code (MCNP). Simulated components include x-ray target, primary conical collimator, x-ray beam flattening filter and secondary collimators. X-ray photon energy spectra and angular distributions were calculated using the model. The x-ray beam emerging from the secondary collimators were scored by considering the total x-ray spectra from the target as the source of x-rays at the target position. The depth dose distribution and dose profiles at different depths and field sizes have been calculated at a nominal operating potential of 6 MV and found to be within acceptable limits. It is concluded that accurate specification of the component dimensions, composition and nominal accelerating potential gives a good assessment of the x-ray energy spectra.

  5. X-ray face mask and chest shield device

    International Nuclear Information System (INIS)

    Moti, S.

    1981-01-01

    A protective face mask is designed to shield an x-ray technician or machine operator primarily from random secondary or scatter x-rays deflected towards his face, head and neck by the table, walls, equipment and other reflecting elements in an x-ray room or chamber. The face mask and chest shield device can be mounted on a patient's shoulders in reverse attitude to protect the back of a patient's head and neck from the x-ray beam. The face mask is relatively or substantially transparent and contains lead in combination with a plastic ionomer or comonomer, which to a degree absorbs or resists penetration of the random deflected secondary or scatter x-rays or the x-ray beam through the mask. The face mask is removably attachable to the chest shield for easy application of the device to and support upon the shoulders of the technician or the patient. (author)

  6. Exposure reduction in general dental practice using digital x-ray imaging system for intraoral radiography with additional x-ray beam filter

    International Nuclear Information System (INIS)

    Shibuya, Hitoshi; Mori, Toshimichi; Hayakawa, Yoshihiko; Kuroyanagi, Kinya; Ota, Yoshiko

    1997-01-01

    To measure exposure reduction in general dental practice using digital x-ray imaging systems for intraoral radiography with additional x-ray beam filter. Two digital x-ray imaging systems, Pana Digital (Pana-Heraus Dental) and CDR (Schick Technologies), were applied for intraoral radiography in general dental practice. Due to the high sensitivity to x-rays, additional x-ray beam filters for output reduction were used for examination. An Orex W II (Osada Electric Industry) x-ray generator was operated at 60 kVp, 7 mA. X-ray output (air-kerma; Gy) necessary for obtaining clinically acceptable images was measured at 0 to 20 cm in 5 cm steps from the cone tip using an ionizing chamber type 660 (Nuclear Associates) and compared with those for Ektaspeed Plus film (Eastman Kodak). The Pana Digital system was used with the optional filter supplied by Pana-Heraus Dental which reduced the output to 38%. The exposure necessary to obtain clinically acceptable images was only 40% of that for the film. The CDR system was used with the Dental X-ray Beam Filter Kit (Eastman Kodak) which reduced the x-ray output to 30%. The exposure necessary to obtain clinically acceptable images was only 20% of that for the film. The two digital x-ray imaging systems, Pana Digital and CDR, provided large dose savings (60-80%) compared with Ektaspeed Plus film when applied for intraoral radiography in general dental practice. (author)

  7. Low energy x-ray spectrometer

    International Nuclear Information System (INIS)

    Woodruff, W.R.

    1981-01-01

    A subkilovolt spectrometer has been produced to permit high-energy-resolution, time-dependent x-ray intensity measurements. The diffracting element is a curved mica (d = 9.95A) crystal. To preclude higher order (n > 1) diffractions, a carbon x-ray mirror that reflects only photons with energies less than approx. 1.1 keV is utilized ahead of the diffracting element. The nominal energy range of interest is 800 to 900 eV. The diffracted photons are detected by a gold-surface photoelectric diode designed to have a very good frequency response, and whose current is recorded on an oscilloscope. A thin, aluminium light barrier is placed between the diffracting crystal and the photoelectric diode detector to keep any uv generated on or scattered by the crystal from illuminating the detector. High spectral energy resolution is provided by many photocathodes between 8- and 50-eV wide placed serially along the diffracted x-ray beam at the detector position. The spectrometer was calibrated for energy and energy dispersion using the Ni Lα 1 2 lines produced in the LLNL IONAC accelerator and in third order using a molybdenum target x-ray tube. For the latter calibration the carbon mirror was replaced by one surfaced with rhodium to raise the cut-off energy to about 3 keV. The carbon mirror reflection dependence on energy was measured using one of our Henke x-ray sources. The curved mica crystal diffraction efficiency was measured on our Low-Energy x-ray (LEX) machine. The spectrometer performs well although some changes in the way the x-ray mirror is held are desirable. 16 figures

  8. SU-E-I-52: Effect of Various X-Ray Beam Qualities On the Exposure Index

    International Nuclear Information System (INIS)

    Yasumatsu, S; Iwase, K; Shimizu, Y; Tanaka, N; Morishita, J

    2015-01-01

    Purpose: The exposure index (EI) proposed by the International Electrotechnical Commission (IEC) 62494-1 is expected to be utilized as a standard dose index by every manufacturer. The IEC recommended the usage of RQA5 for the EI. However, X-ray beam qualities, particularly in clinical practices, vary depending on the examination objects and exposure conditions, including usage of anti-scatter grids. We investigated the effects of the X-ray beam qualities other than RQA5 on the EI. Methods: The Xray beam qualities of RQA3, 5, 7, and 9 in IEC 61267 Ed. 1.0 were adopted in a computed radiography system. A uniform exposure without objects was performed to measure the exposure indicators (S values) and air kerma (K). The relational equations between the S values and K were derived for the determination of the EI values. The EI values for RQA3, 7, and 9 were compared to those for RQA5 at the fixed S values of 100, 200, 400, and 600. Finally, the half-value layers (HVLs) using four grids (ratio 6:1, 8:1, 10:1, and 12:1) for the RQA5 X-ray were compared to those with RQA3–9. Results: The EI values for RQA3, 7, and 9 were up to 35.3%, 11.8%, and 38.7% higher, respectively, than that for RQA5 at the S value of 600. The HVLs without grids and with various grids for RQA5 were 6.85 mm Al. and in the range of 6.94–7.29 mm Al. (ΔHVL: up to 0.44 mm Al.), respectively. This variation in the HVLs with grids was smaller than that observed for RQA3–9 (ΔHVL: 2.0–7.5 mm Al.). Conclusion: Although the usage of grids may not greatly affect the EI, the X-ray beam quality for the determination of the EI cannot be ignored in the clinical evaluation of the dose index

  9. New X-ray beam position monitors with submicron resolution utilizing imaging of scattered X-rays at CHESS

    International Nuclear Information System (INIS)

    Revesz, Peter; Temnykh, Alexander B.; Pauling, Alan K.

    2011-01-01

    At CHESS' A, F and G wiggler beam lines three new video beam position monitors (VBPMs) have been commissioned. These new VBPMs utilize X-rays scattered from the graphite filter (A and F line) or from a beryllium window (G-line) as the white wiggler beam passes through them. As the X-rays scatter in all directions from the scattering medium, a slit camera creates an image of the beam's footprint on a fluorescent screen. This image is then viewed by a CCD camera and analyzed using a computer program to calculate the intensity centroid, the beam profile and integrated intensity. These data are delivered to the CHESS signal archiving system for storage and display. The new systems employ digital cameras. These cameras are free of the noise inherent to the analog systems with long video signal connections. As a result, the beam position data delivered by the new systems are more reliable and accurate as shown by beam position traces using different beam position monitors on the same beam line.

  10. New X-ray beam position monitors with submicron resolution utilizing imaging of scattered X-rays at CHESS

    Energy Technology Data Exchange (ETDEWEB)

    Revesz, Peter, E-mail: pr20@cornell.edu [Cornell University, Cornell High Energy Synchrotron Source, Ithaca 14850, NY (United States); Temnykh, Alexander B. [Cornell University, Laboratory for Elem-Particle Physics, Ithaca 14850, NY (United States); Pauling, Alan K. [Cornell University, Cornell High Energy Synchrotron Source, Ithaca 14850, NY (United States)

    2011-09-01

    At CHESS' A, F and G wiggler beam lines three new video beam position monitors (VBPMs) have been commissioned. These new VBPMs utilize X-rays scattered from the graphite filter (A and F line) or from a beryllium window (G-line) as the white wiggler beam passes through them. As the X-rays scatter in all directions from the scattering medium, a slit camera creates an image of the beam's footprint on a fluorescent screen. This image is then viewed by a CCD camera and analyzed using a computer program to calculate the intensity centroid, the beam profile and integrated intensity. These data are delivered to the CHESS signal archiving system for storage and display. The new systems employ digital cameras. These cameras are free of the noise inherent to the analog systems with long video signal connections. As a result, the beam position data delivered by the new systems are more reliable and accurate as shown by beam position traces using different beam position monitors on the same beam line.

  11. Flat Field Anomalies in an X-Ray CCD Camera Measured Using a Manson X-Ray Source

    International Nuclear Information System (INIS)

    Michael Haugh

    2008-01-01

    The Static X-ray Imager (SXI) is a diagnostic used at the National Ignition Facility (NIF) to measure the position of the X-rays produced by lasers hitting a gold foil target. It determines how accurately NIF can point the laser beams and is critical to proper NIF operation. Imagers are located at the top and the bottom of the NIF target chamber. The CCD chip is an X-ray sensitive silicon sensor, with a large format array (2k x 2k), 24 (micro)m square pixels, and 15 (micro)m thick. A multi-anode Manson X-ray source, operating up to 10kV and 2mA, was used to characterize and calibrate the imagers. The output beam is heavily filtered to narrow the spectral beam width, giving a typical resolution E/ΔE ∼ 12. The X-ray beam intensity was measured using an absolute photodiode that has accuracy better than 1% up to the Si K edge and better than 5% at higher energies. The X-ray beam provides full CCD illumination and is flat, within ±1.5% maximum to minimum. The spectral efficiency was measured at 10 energy bands ranging from 930 eV to 8470 eV. The efficiency pattern follows the properties of Si. The maximum quantum efficiency is 0.71. We observed an energy dependent pixel sensitivity variation that showed continuous change over a large portion of the CCD. The maximum sensitivity variation was >8% at 8470 eV. The geometric pattern did not change at lower energies, but the maximum contrast decreased and was less than the measurement uncertainty below 4 keV. We were also able to observe debris on the CCD chip. The debris showed maximum contrast at the lowest energy used, 930 eV, and disappeared by 4 keV. The Manson source is a powerful tool for characterizing the imaging errors of an X-ray CCD imager. These errors are quite different from those found in a visible CCD imager

  12. Verification of industrial x-ray machine: MINTs experience

    International Nuclear Information System (INIS)

    Aziz Amat; Saidi Rajab; Eesan Pasupathi; Saipo Bahari Abdul Ratan; Shaharudin Sayuti; Abd Nassir Ibrahim; Abd Razak Hamzah

    2005-01-01

    Radiation and electrical safety of the industrial x-ray equipment required to meet Atomic Energy Licensing Board(AELB) guidelines ( LEM/TEK/42 ) at the time of installation and subsequently a periodic verification should be ensured. The purpose of the guide is to explain the requirements employed in conducting the test on industrial x-ray apparatus and be certified in meeting with our local legislative and regulation. Verification is aimed to provide safety assurance information on electrical requirements and the minimum radiation exposure to the operator. This regulation is introduced on new models imported into the Malaysian market. Since June, 1997, Malaysian Institute for Nuclear Technology Research (MINT) has been approved by AELB to provide verification services to private company, government and corporate body throughout Malaysia. Early January 1997, AELB has made it mandatory that all x-ray equipment for industrial purpose (especially Industrial Radiography) must fulfill certain performance test based on the LEM/TEK/42 guidelines. MINT as the third party verification encourages user to improve maintenance of the equipment. MINT experiences in measuring the performance on intermittent and continuous duty rating single-phase industrial x-ray machine in the year 2004 indicated that all of irradiating apparatus tested pass the test and met the requirements of the guideline. From MINT record, 1997 to 2005 , three x-ray models did not meet the requirement and thus not allowed to be used unless the manufacturers willing to modify it to meet AELB requirement. This verification procedures on electrical and radiation safety on industrial x-ray has significantly improved the the maintenance cultures and safety awareness in the usage of x-ray apparatus in the industrial environment. (Author)

  13. Beam synchronous detection techniques for X-Ray spectroscopy

    International Nuclear Information System (INIS)

    Goujon, Gérard; Rogalev, Andreï; Goulon, José; Feite, Serge; Wilhelm, Fabrice

    2013-01-01

    The Photo diode detectors combine a set of properties that make them most appropriate, in particular, for X-ray Magnetic Circular Dichroism (XMCD) experiments. Under standard operating conditions, the detection bandwidth is primarily limited by the transimpedance preamplifier that converts the very low ac photocurrent into a voltage. On the other hand, when the photodiode is reverse biased, its finite shunt resistance will cause an undesirable, temperature dependent DC dark current. The best strategy to get rid of it is to use synchronous detection techniques. A classical implementation is based on the use of a chopper modulating the X-ray beam intensity at rather low frequencies (typically below 1 kHz). Here we report on the recent development of a fast Xray detection which has the capability to fully exploit the frequency structure of the ESRF X-ray beam (355 KHz and its harmonics). The availability of new wide band preamplifiers allowed us to extend the working frequency range up to a few MHz. A beam synchronous data processing was implemented in large FPGAs. Performances of the new detection system implemented at the ESRF beamline ID12 are illustrated with detection of the Fe K-edge XMCD spectra in garnets, using 4 bunches operation mode with modulation frequency of 1.4 MHz.

  14. Determination of the effective energy in X-rays standard beams, mammography level

    Energy Technology Data Exchange (ETDEWEB)

    Correa, Eduardo de Lima; Vivolo, Vitor; Potiens, Maria da Penha A., E-mail: Vivolo@ipen.b, E-mail: mppalbu@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2011-07-01

    The X-rays beams used in diagnostic radiology are heterogeneous. This means that, in a radiological beam, it can be found photons with different energies. Because of that is common to work with the concept of effective energy. In this study the effective energy of an X-rays system used in instruments calibration was determined, as part of the mammography radiation qualities establishment. The procedure presented here was developed based on information found in the literature. The X-ray mass attenuation coefficients for aluminum, given by NIST web site, were used and the mathematical adjusts were done using the Origin 8.0 program. The results are part of the mammographic X-rays beams characteristics determination and it is important to keep the quality of this reference system. (author)

  15. X-ray luminescence computed tomography imaging via multiple intensity weighted narrow beam irradiation

    Science.gov (United States)

    Feng, Bo; Gao, Feng; Zhao, Huijuan; Zhang, Limin; Li, Jiao; Zhou, Zhongxing

    2018-02-01

    The purpose of this work is to introduce and study a novel x-ray beam irradiation pattern for X-ray Luminescence Computed Tomography (XLCT), termed multiple intensity-weighted narrow-beam irradiation. The proposed XLCT imaging method is studied through simulations of x-ray and diffuse lights propagation. The emitted optical photons from X-ray excitable nanophosphors were collected by optical fiber bundles from the right-side surface of the phantom. The implementation of image reconstruction is based on the simulated measurements from 6 or 12 angular projections in terms of 3 or 5 x-ray beams scanning mode. The proposed XLCT imaging method is compared against the constant intensity weighted narrow-beam XLCT. From the reconstructed XLCT images, we found that the Dice similarity and quantitative ratio of targets have a certain degree of improvement. The results demonstrated that the proposed method can offer simultaneously high image quality and fast image acquisition.

  16. 1- to 10-keV x-ray backlighting of annular wire arrays on the Sandia Z-machine using bent-crystal imaging techniques

    International Nuclear Information System (INIS)

    Rambo, Patrick K.; Wenger, David Franklin; Bennett, Guy R.; Sinars, Daniel Brian; Smith, Ian Craig; Porter, John Larry Jr.; Cuneo, Michael Edward; Rovang, Dean Curtis; Anderson, Jessica E.

    2003-01-01

    Annular wire array implosions on the Sandia Z-machine can produce >200 TW and 1-2 MJ of soft x rays in the 0.1-10 keV range. The x-ray flux and debris in this environment present significant challenges for radiographic diagnostics. X-ray backlighting diagnostics at 1865 and 6181 eV using spherically-bent crystals have been fielded on the Z-machine, each with a ∼0.6 eVspectral bandpass, 10 (micro)m spatial resolution, and a 4 mm by 20mm field of view. The Z-Beamlet laser, a 2-TW, 2-kJ Nd:glass laser(λ = 527 nm), is used to produce 0.1-1 J x-ray sources for radiography. The design, calibration, and performance of these diagnostics is presented.

  17. Investigation of a novel x-ray tube for the calibration of the x-ray crystal spectrometer in the KSTAR machine

    International Nuclear Information System (INIS)

    Bak, J.G.; Lee, S.G.

    2007-01-01

    A novel x-ray tube with a line filament has been developed for the in-situ calibration of the x-ray crystal spectrometer (XCS) in the KSTAR machine. The characteristics of the x-ray tube are investigated from the x-ray images obtained by using a pinhole and a CCD detector. It is found that the image has the width of about 0.1 mm, which is much improved as compared with the previous experimental results. In addition, there is a uniform region around the center of the image within its full length of 13.5 mm. This work may lead to the development of a novel x-ray tube with a line focus, which is required for the calibration of the XCS. Experimental results from the investigation of the x-ray tube are presented and the technical issues in a design of the in-situ calibration system using the x-ray tube for the KSTAR XCS are discussed. (author)

  18. Soft X-ray beam induced current technique

    Energy Technology Data Exchange (ETDEWEB)

    Watts, B; Ade, H [Department of Physics, North Carolina State University, Raleigh, NC 27695 (United States); Queen, D; Hellman, F [Department of Physics, University of California, Berkeley, CA 94720 (United States); Kilcoyne, A L D; Tyliszczak, T, E-mail: benjamin.watts@gmail.co [Advanced Light Source, Lawrence Berkeley Nat. Lab., Berkeley, CA 94720 (United States)

    2009-09-01

    Direct mapping of the charge transport efficiency of polymer solar cell devices using a soft X-ray beam induced current (SoXBIC) method is described. By fabricating a polymer solar cell on an x-ray transparent substrate, we demonstrate the ability to map polymer composition and nanoscale structure within an operating solar cell device and to simultaneously measure the local charge transport efficiency via the short-circuit current. A simple model is calculated and compared to experimental SoXBIC data of a PFB:F8BT bulk-heterojunction device in order to gain greater insight into the device operation and physics.

  19. Comparison of neutron and high-energy X-ray dual-beam radiography for air cargo inspection

    International Nuclear Information System (INIS)

    Liu, Y.; Sowerby, B.D.; Tickner, J.R.

    2008-01-01

    Dual-beam radiography techniques utilising various combinations of high-energy X-rays and neutrons are attractive for screening bulk cargo for contraband such as narcotics and explosives. Dual-beam radiography is an important enhancement to conventional single-beam X-ray radiography systems in that it provides additional information on the composition of the object being imaged. By comparing the attenuations of transmitted dual high-energy beams, it is possible to build a 2D image, colour coded to indicate material. Only high-energy X-rays, gamma-rays and neutrons have the required penetration to screen cargo containers. This paper reviews recent developments and applications of dual-beam radiography for air cargo inspection. These developments include dual high-energy X-ray techniques as well as fast neutron and gamma-ray (or X-ray) radiography systems. High-energy X-ray systems have the advantage of generally better penetration than neutron systems, depending on the material being interrogated. However, neutron systems have the advantage of much better sensitivity to material composition compared to dual high-energy X-ray techniques. In particular, fast neutron radiography offers the potential to discriminate between various classes of organic material, unlike dual energy X-ray techniques that realistically only offer the ability to discriminate between organic and metal objects

  20. Highly porous nanoberyllium for X-ray beam speckle suppression

    Energy Technology Data Exchange (ETDEWEB)

    Goikhman, Alexander, E-mail: agoikhman@ymail.com; Lyatun, Ivan; Ershov, Petr [Immanuel Kant Baltic Federal University, Nevskogo str. 14, Kaliningrad 236041 (Russian Federation); Snigireva, Irina [European Synchrotron Radiation Facility, BP 220, 38043 Grenoble (France); Wojda, Pawel [Immanuel Kant Baltic Federal University, Nevskogo str. 14, Kaliningrad 236041 (Russian Federation); Gdańsk University of Technology, 11/12 G. Narutowicza, Gdańsk 80-233 (Poland); Gorlevsky, Vladimir; Semenov, Alexander; Sheverdyaev, Maksim; Koletskiy, Viktor [A. A. Bochvar High-Technology Scientific Research Institute for Inorganic Materials, Rogova str. 5a, Moscow 123098 (Russian Federation); Snigirev, Anatoly [Immanuel Kant Baltic Federal University, Nevskogo str. 14, Kaliningrad 236041 (Russian Federation); European Synchrotron Radiation Facility, BP 220, 38043 Grenoble (France)

    2015-04-09

    A speckle suppression device containing highly porous nanoberyllium is proposed for manipulating the spatial coherence length and removing undesirable speckle structure during imaging experiments. This paper reports a special device called a ‘speckle suppressor’, which contains a highly porous nanoberyllium plate squeezed between two beryllium windows. The insertion of the speckle suppressor in an X-ray beam allows manipulation of the spatial coherence length, thus changing the effective source size and removing the undesirable speckle structure in X-ray imaging experiments almost without beam attenuation. The absorption of the nanoberyllium plate is below 1% for 1 mm thickness at 12 keV. The speckle suppressor was tested on the ID06 ESRF beamline with X-rays in the energy range from 9 to 15 keV. It was applied for the transformation of the phase–amplitude contrast to the pure amplitude contrast in full-field microscopy.

  1. Highly porous nanoberyllium for X-ray beam speckle suppression

    International Nuclear Information System (INIS)

    Goikhman, Alexander; Lyatun, Ivan; Ershov, Petr; Snigireva, Irina; Wojda, Pawel; Gorlevsky, Vladimir; Semenov, Alexander; Sheverdyaev, Maksim; Koletskiy, Viktor; Snigirev, Anatoly

    2015-01-01

    A speckle suppression device containing highly porous nanoberyllium is proposed for manipulating the spatial coherence length and removing undesirable speckle structure during imaging experiments. This paper reports a special device called a ‘speckle suppressor’, which contains a highly porous nanoberyllium plate squeezed between two beryllium windows. The insertion of the speckle suppressor in an X-ray beam allows manipulation of the spatial coherence length, thus changing the effective source size and removing the undesirable speckle structure in X-ray imaging experiments almost without beam attenuation. The absorption of the nanoberyllium plate is below 1% for 1 mm thickness at 12 keV. The speckle suppressor was tested on the ID06 ESRF beamline with X-rays in the energy range from 9 to 15 keV. It was applied for the transformation of the phase–amplitude contrast to the pure amplitude contrast in full-field microscopy

  2. Cone Beam X-Ray Luminescence Tomography Imaging Based on KA-FEM Method for Small Animals.

    Science.gov (United States)

    Chen, Dongmei; Meng, Fanzhen; Zhao, Fengjun; Xu, Cao

    2016-01-01

    Cone beam X-ray luminescence tomography can realize fast X-ray luminescence tomography imaging with relatively low scanning time compared with narrow beam X-ray luminescence tomography. However, cone beam X-ray luminescence tomography suffers from an ill-posed reconstruction problem. First, the feasibility of experiments with different penetration and multispectra in small animal has been tested using nanophosphor material. Then, the hybrid reconstruction algorithm with KA-FEM method has been applied in cone beam X-ray luminescence tomography for small animals to overcome the ill-posed reconstruction problem, whose advantage and property have been demonstrated in fluorescence tomography imaging. The in vivo mouse experiment proved the feasibility of the proposed method.

  3. Thermoluminescence of zirconium oxide nanostructured to mammography X-ray beams

    International Nuclear Information System (INIS)

    Palacios, L.L.; Rivera, T.; Roman, J.; Azorín, J.; Gaona, E.

    2012-01-01

    In the present work thermoluminescent (TL) response of zirconium oxide (ZrO 2 ) nanostructured induced by mammography X-ray radiation was investigated. Measurements were made of the response per unit air kerma of ZrO 2 with mammography equipment parameters (semiautomatic exposure control, 24 kVp and 108 mAs). The calibration curves were obtained by simultaneously irradiating ZrO 2 samples and ion chamber. Samples of ZrO 2 showed a linear response as a function of entrance skin air kerma. The observed results in TL properties suggest that ZrO 2 nanostructured could be considered as an effective material for X-ray beams dosimetry if appropriate calibration procedures are performed. - Highlights: ► X-ray low energy thermoluminescent of ZrO 2 dosimeter is developed. ► Air kerma measurements were made by thermoluminescent dosimeter ZrO 2 using mammography equipment parameters. ► Entrance surface skin doses were made using thermoluminescent dosimeter of ZrO 2 to X-ray beam quality control.

  4. Development and commissioning of an x-ray beam alignment flag for NSLS-II

    Energy Technology Data Exchange (ETDEWEB)

    Kosciuk, B., E-mail: bkosciuk@bnl.gov; Hu, Y.; Keister, J.; Seletskiy, S. [National Synchrotron Light Source-II, Brookhaven National Laboratory, Upton, New York 11973 (United States)

    2016-07-27

    The NSLS-II Synchrotron Light Source is a 3 GeV electron storage ring recently commissioned and is now entering operations at Brookhaven National Laboratory. One of the major tasks was to commission the six project beamline front ends which required a diagnostic to resolve x-ray beam position for the purpose of beam alignment at low current. Since none of the front ends were outfitted with any x-ray diagnostics in the baseline design, an x-ray beam profile monitor or “flag” that could be easily installed into existing front end vacuum chambers was proposed to satisfy this requirement. Here we present the development of this novel device which utilizes a polycrystalline CVD diamond luminescent screen to produce a visible image of the x-ray beam cross-section and is then captured with a CCD camera.

  5. Micro-beam X-ray fluorescence and absorption imaging techniques at the IAEA Laboratories

    International Nuclear Information System (INIS)

    Wegrzynek, Dariusz; Markowicz, A.; Bamford, S.; Chinea-Cano, E.; Bogovac, M.

    2005-01-01

    X-ray tube based, micro-beam X-ray fluorescence scanning spectrometer has been equipped with two energy dispersive X-ray detectors. The two-detector configuration allows for simultaneous collection of X-ray fluorescence (XRF) and transmitted X-ray beam signals with a spatial resolution in the range of 10-50 μm, depending on the X-ray focussing element in use. The XRF signal is collected with a standard, liquid nitrogen cooled Si(Li) detector. The X-ray beam transmitted through the sample is acquired with a thermoelectrically cooled, silicon drift (SD) detector. The data acquisition is carried out in a fully automatic way under control of the SPECTOR-LOCATOR software. The software controls the scanning procedure and X-ray spectra acquisition during the scan. The energy dispersive X-ray spectra collected at every 'pixel' are stored for off-line processing. For selected regions of interest (ROI's), the element maps are constructed and displayed on-line. The spectrometer has been used for mapping elemental distributions and for performing 2D- and 3D-tomograpic imaging of minute objects in X-ray absorption and in X-ray fluorescence mode. A unique feature of the described system is simultaneous utilization of the two detectors, Si(Li) and SD, which adds new options for quantitative analysis and data interpretation. Examples of elemental mapping and 3D tomographic imaging as well as the advanced features of the SPECTOR-LOCATOR measurement control and data acquisition software are presented in this work

  6. Influence of beam divergence on form-factor in X-ray diffraction radiation

    International Nuclear Information System (INIS)

    Sergeeva, D.Yu.; Tishchenko, A.A.; Strikhanov, M.N.

    2015-01-01

    Diffraction radiation from divergent beam is considered in terms of radiation in UV and X-ray range. Scedastic form of Gaussian distribution of the particle in the bunch, i.e. Gaussian distribution with changing dispersion has been used, which is more adequate for description of divergent beams than often used Gaussian distribution with constant dispersion. Both coherent and incoherent form-factors are taken into account. The conical diffraction effect in diffraction radiation is proved to make essential contribution in spectral-angular characteristics of radiation from a divergent beam

  7. MIT modular x-ray source systems for the study of plasma diagnostics

    Science.gov (United States)

    Coleman, J. W.; Wenzel, K. W.; Petrasso, R. D.; Lo, D. H.; Li, C. K.; Lierzer, J. R.; Wei, T.

    1992-10-01

    Two new x-ray source systems are now on line at our facility. Each provides an e-beam to 25 kV. Targets are interchangeable between machines, and four x-ray detectors may be used simultaneously with a target. The gridded e-gun of the RACEHORSE system gives a 0.5-1.0-cm pulsable spot on target. The nongridded e-gun of the SCORPION system provides a 0.3-mm or smaller dc microspot on target. RACEHORSE is being used to study and characterize type-II diamond photoconductors for use in diagnosing plasmas, while SCORPION is being used to develop a slitless spectrograph using photographic film. Source design details and some RACEHORSE results are presented.

  8. Characterization and control of femtosecond electron and X-ray beams at free-electron lasers

    International Nuclear Information System (INIS)

    Behrens, Christopher

    2012-11-01

    X-ray free-electron lasers (FELs) open up new frontiers in photon science, and in order to take full advantage of these unique accelerator-based light sources, the characterization and control of the femtosecond electron and X-ray beams is essential. Within this cumulative thesis, recent results achieved within the active research field of femtosecond electron and X-ray beams at FELs are reported.The basic principles of X-ray FELs are described, and concepts of longitudinal electron beam diagnostics with femtosecond accuracy are covered. Experimental results obtained with a transverse deflecting structure (TDS) and spectroscopy of coherent terahertz radiation are presented, and the suppression of coherent optical radiation effects, required for diagnostics utilizing a TDS, is demonstrated. Control of the longitudinal phase space by using multiple radio frequencies for longitudinal electron beam tailoring is presented, and a new technique of reversible electron beam heating with two TDSs is described. For the characterization of femtosecond X-ray pulses, a novel method based on dedicated longitudinal phase space diagnostics for electron beams is introduced, and recent measurements with a streaking technique using external terahertz fields are presented.

  9. Characterization and control of femtosecond electron and X-ray beams at free-electron lasers

    Energy Technology Data Exchange (ETDEWEB)

    Behrens, Christopher

    2012-11-15

    X-ray free-electron lasers (FELs) open up new frontiers in photon science, and in order to take full advantage of these unique accelerator-based light sources, the characterization and control of the femtosecond electron and X-ray beams is essential. Within this cumulative thesis, recent results achieved within the active research field of femtosecond electron and X-ray beams at FELs are reported.The basic principles of X-ray FELs are described, and concepts of longitudinal electron beam diagnostics with femtosecond accuracy are covered. Experimental results obtained with a transverse deflecting structure (TDS) and spectroscopy of coherent terahertz radiation are presented, and the suppression of coherent optical radiation effects, required for diagnostics utilizing a TDS, is demonstrated. Control of the longitudinal phase space by using multiple radio frequencies for longitudinal electron beam tailoring is presented, and a new technique of reversible electron beam heating with two TDSs is described. For the characterization of femtosecond X-ray pulses, a novel method based on dedicated longitudinal phase space diagnostics for electron beams is introduced, and recent measurements with a streaking technique using external terahertz fields are presented.

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... of knee x-rays. A portable x-ray machine is a compact apparatus that can be taken ... of the body being examined, an x-ray machine produces a small burst of radiation that passes ...

  11. A new high-speed X-ray beam chopper

    International Nuclear Information System (INIS)

    McPherson, A.; Wang, J.; Lee, P. L.; Mills, D. M.

    1999-01-01

    A new high-speed x-ray beam chopper using laser scanner technology has been developed and tested on the SRI-CAT sector 1 beamline at the Advanced Photon Source (APS) storage ring (1). As illustrated in figure 1, it is compact in size and has two sets of transmission windows: BK-7 glass for visible light transmission and 0.23-mm-thick Be for the transmission of x-rays. The rotor is made of aluminum and has a diameter of 50.8 mm. A 0.5-mm-wide and 2.29-mm-tall slit is cut through the center of the rotor. The circumference of the rotor has a coating of 1-mm-thick Ni, which gives an attenuation of 10 8 at 30 keV. Turning at nearly 80000 RPM, this beam chopper has an opening time window of 2450 ns, corresponding to 67% of the revolution time of the APS storage ring. The primary feature in selecting laser scanner technology to develop into an x-ray beam chopper was the high level of rotational speed control of the rotor that makes up the beam chopper element (2). By using an optical feedback circuit to sample the rotational speed four times each revolution, the jitter in the position of the transmission open time window is only 3 ns at the 3 standard deviation level. The APS storage ring orbital frequency, supplied by the control room, is divided down to provide the appropriate drive frequency for the beam chopper motor controller. By this means, both the storage ring and the beam chopper are operating off the same master clock. After a turn-on time of about 15 to 20 seconds, the rotational precision of the motor results in immediate phase locking to the temporal structure of the APS storage ring. By inserting a Stanford delay generator between the frequency divider and the beam chopper motor controller, the phase between the storage ring temporal structure and the beam chopper rotation can be adjusted to position the transmission time window of the beam chopper on any desired part of the storage ring fill pattern. If an asymmetric fill pattern is used in the APS storage

  12. Thermoluminescence of zirconium oxide nanostructured to mammography X-ray beams

    Energy Technology Data Exchange (ETDEWEB)

    Palacios, L.L. [Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada-Legaria, IPN. Av. Legaria 694, 11500 Mexico D.F. (Mexico); Rivera, T., E-mail: trivera@ipn.mx [Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada-Legaria, IPN. Av. Legaria 694, 11500 Mexico D.F. (Mexico); Roman, J. [Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada-Legaria, IPN. Av. Legaria 694, 11500 Mexico D.F. (Mexico); Azorin, J. [Universidad Autonoma Metropolitana-Iztapalapa. Av. San Rafael Atlixco 187, 09340 Mexico D.F. (Mexico); Gaona, E. [Universidad Autonoma Metropolitana-Xochimilco. Calz. Del Hueso 1100, 04960 Mexico D.F. (Mexico)

    2012-07-15

    In the present work thermoluminescent (TL) response of zirconium oxide (ZrO{sub 2}) nanostructured induced by mammography X-ray radiation was investigated. Measurements were made of the response per unit air kerma of ZrO{sub 2} with mammography equipment parameters (semiautomatic exposure control, 24 kVp and 108 mAs). The calibration curves were obtained by simultaneously irradiating ZrO{sub 2} samples and ion chamber. Samples of ZrO{sub 2} showed a linear response as a function of entrance skin air kerma. The observed results in TL properties suggest that ZrO{sub 2} nanostructured could be considered as an effective material for X-ray beams dosimetry if appropriate calibration procedures are performed. - Highlights: Black-Right-Pointing-Pointer X-ray low energy thermoluminescent of ZrO{sub 2} dosimeter is developed. Black-Right-Pointing-Pointer Air kerma measurements were made by thermoluminescent dosimeter ZrO{sub 2} using mammography equipment parameters. Black-Right-Pointing-Pointer Entrance surface skin doses were made using thermoluminescent dosimeter of ZrO{sub 2} to X-ray beam quality control.

  13. X-ray pencil beam facility for optics characterization

    Science.gov (United States)

    Krumrey, Michael; Cibik, Levent; Müller, Peter; Bavdaz, Marcos; Wille, Eric; Ackermann, Marcelo; Collon, Maximilien J.

    2010-07-01

    The Physikalisch-Technische Bundesanstalt (PTB) has used synchrotron radiation for the characterization of optics and detectors for astrophysical X-ray telescopes for more than 20 years. At a dedicated beamline at BESSY II, a monochromatic pencil beam is used by ESA and cosine Research since the end of 2005 for the characterization of novel silicon pore optics, currently under development for the International X-ray Observatory (IXO). At this beamline, a photon energy of 2.8 keV is selected by a Si channel-cut monochromator. Two apertures at distances of 12.2 m and 30.5 m from the dipole source form a pencil beam with a typical diameter of 100 μm and a divergence below 1". The optics to be investigated is placed in a vacuum chamber on a hexapod, the angular positioning is controlled by means of autocollimators to below 1". The reflected beam is registered at 5 m distance from the optics with a CCD-based camera system. This contribution presents design and performance of the upgrade of this beamline to cope with the updated design for IXO. The distance between optics and detector can now be 20 m. For double reflection from an X-ray Optical Unit (XOU) and incidence angles up to 1.4°, this corresponds to a vertical translation of the camera by 2 m. To achieve high reflectance at this angle even with uncoated silicon, a lower photon energy of 1 keV is available from a pair of W/B4C multilayers. For coated optics, a high energy option can provide a pencil beam of 7.6 keV radiation.

  14. A water-cooled x-ray monochromator for using off-axis undulator beam

    International Nuclear Information System (INIS)

    Khounsary, A.; Maser, J.

    2000-01-01

    Undulator beamlines at third-generation synchrotrons x-ray sources are designed to use the high-brilliance radiation that is contained in the central cone of the generated x-ray beams. The rest of the x-ray beam is often unused. Moreover, in some cases, such as in the zone-plate-based microfocusing beamlines, only a small part of the central radiation cone around the optical axis is used. In this paper, a side-station branch line at the Advanced Photon Source that takes advantage of some of the unused off-axis photons in a microfocusing x-ray beamline is described. Detailed information on the design and analysis of a high-heat-load water-cooled monochromator developed for this beamline is provided

  15. Investigation of chemical vapour deposition diamond detectors by X-ray micro-beam induced current and X-ray micro-beam induced luminescence techniques

    International Nuclear Information System (INIS)

    Olivero, P.; Manfredotti, C.; Vittone, E.; Fizzotti, F.; Paolini, C.; Lo Giudice, A.; Barrett, R.; Tucoulou, R.

    2004-01-01

    Tracking detectors have become an important ingredient in high-energy physics experiments. In order to survive the harsh detection environment of the large hadron collider (LHC), trackers need to have special properties. They must be radiation hard, provide fast collection of charge, be as thin as possible and remove heat from readout electronics. The unique properties of diamond allow it to fulfill these requirements. In this work we present an investigation of the charge transport and luminescence properties of 'detector grade' artificial chemical vapour deposition (CVD) diamond devices developed within the CERN RD42 collaboration, performed by means of X-ray micro-beam induced current collection (XBICC) and X-ray micro-beam induced luminescence (XBIL) techniques. XBICC technique allows quantitative estimates of the transport parameters of the material to be evaluated and mapped with micrometric spatial resolution. In particular, the high resolution and sensitivity of the technique has allowed a quantitative study of the inhomogeneity of the charge transport parameter defined as the product of mobility and lifetime for both electron and holes. XBIL represents a technique complementary to ion beam induced luminescence (IBIL), which has already been used by our group, since X-ray energy loss profile in the material is different from that of MeV ions. X-ray induced luminescence maps have been performed simultaneously with induced photocurrent maps, to correlate charge transport and induced luminescence properties of diamond. Simultaneous XBICC and XBIL maps exhibit features of partial complementarity that have been interpreted on the basis of considerations on radiative and non-radiative recombination processes which compete with charge transport efficiency

  16. A fax-machine amorphous silicon sensor for X-ray detection

    Energy Technology Data Exchange (ETDEWEB)

    Alberdi, J. [Association EURATOM/CIEMAT, Madrid (Spain); Barcala, J.M. [Association EURATOM/CIEMAT, Madrid (Spain); Chvatchkine, V. [Association EURATOM/CIEMAT, Madrid (Spain); Ioudine, I. [Association EURATOM/CIEMAT, Madrid (Spain); Molinero, A. [Association EURATOM/CIEMAT, Madrid (Spain); Navarrete, J.J. [Association EURATOM/CIEMAT, Madrid (Spain); Yuste, C. [Association EURATOM/CIEMAT, Madrid (Spain)

    1996-10-01

    Amorphous silicon detectors have been used, basically, as solar cells for energetics applications. As light detectors, linear sensors are used in fax and photocopier machines because they can be built with a large size, low price and have a high radiation hardness. Due to these performances, amorphous silicon detectors have been used as radiation detectors, and, presently, some groups are developing matrix amorphous silicon detectors with built-in electronics for medical X-ray applications. Our group has been working on the design and development of an X-ray image system based on a commercial fax linear amorphous silicon detector. The sensor scans the selected area and detects light produced by the X-ray in a scintillator placed on the sensor. Image-processing software produces a final image with better resolution and definition. (orig.).

  17. Measurements of fast electron beams and soft X-ray emission from plasma-focus experiments

    Directory of Open Access Journals (Sweden)

    Surała Władysław

    2016-06-01

    Full Text Available The paper reports results of the recent experimental studies of pulsed electron beams and soft X-rays in plasma-focus (PF experiments carried out within a modified PF-360U facility at the NCBJ, Poland. Particular attention was focused on time-resolved measurements of the fast electron beams by means of two different magnetic analyzers, which could record electrons of energy ranging from about 41 keV to about 715 keV in several (6 or 8 measuring channels. For discharges performed with the pure deuterium filling, many strong electron signals were recorded in all the measuring channels. Those signals were well correlated with the first hard X-ray pulse detected by an external scintillation neutron-counter. In some of the analyzer channels, electron spikes (lasting about dozens of nanoseconds and appearing in different instants after the current peculiarity (so-called current dip were also recorded. For several discharges, fast ion beams, which were emitted along the z-axis and recorded with nuclear track detectors, were also investigated. Those measurements confirmed a multibeam character of the ion emission. The time-integrated soft X-ray images, which were taken side-on by means of a pinhole camera and sensitive X-ray films, showed the appearance of some filamentary structures and so-called hot spots. The application of small amounts of admixtures of different heavy noble gases, i.e. of argon (4.8% volumetric, krypton (1.6% volumetric, or xenon (0.8% volumetric, decreased intensity of the recorded electron beams, but increased intensity of the soft X-ray emission and showed more distinct and numerous hot spots. The recorded electron spikes have been explained as signals produced by quasi-mono-energetic microbeams emitted from tiny sources (probably plasma diodes, which can be formed near the observed hot spots.

  18. The histories of capillary optics for x-rays and ion beams in Russia, USA, and Japan

    International Nuclear Information System (INIS)

    Umezawa, Kenji

    2009-01-01

    This article introduces the history of X-ray lens and the present situation of ion beam focusing with glass capillaries systems. The basic technology of X-ray lens using glass capillaries was independently developed over 20 years by Prof. Kumakhov in the former Soviet Union and Dr. Soejima in Japan, respectively. In the 1990's, Prof. W.M. Gibson and his coworkers intensively studied X-rays and neutron optics in Albany, NY, USA. X-ray optics with glass capillaries, in these days is well known in the world. This unique technique was fabricated to collimate X-rays. Also, new ion beam analysis technique with glass capillaries systems has been intensively developed by Dr. Nebiki and Prof. Narusawa in Kochi, Japan. These X-rays and ion beams techniques have brought new application for many fields; X-ray detector, X-ray lithography, X-ray astronomy, microdiffraction, medical therapy and biological applications. (author)

  19. Cone Beam X-ray Luminescence Computed Tomography Based on Bayesian Method.

    Science.gov (United States)

    Zhang, Guanglei; Liu, Fei; Liu, Jie; Luo, Jianwen; Xie, Yaoqin; Bai, Jing; Xing, Lei

    2017-01-01

    X-ray luminescence computed tomography (XLCT), which aims to achieve molecular and functional imaging by X-rays, has recently been proposed as a new imaging modality. Combining the principles of X-ray excitation of luminescence-based probes and optical signal detection, XLCT naturally fuses functional and anatomical images and provides complementary information for a wide range of applications in biomedical research. In order to improve the data acquisition efficiency of previously developed narrow-beam XLCT, a cone beam XLCT (CB-XLCT) mode is adopted here to take advantage of the useful geometric features of cone beam excitation. Practically, a major hurdle in using cone beam X-ray for XLCT is that the inverse problem here is seriously ill-conditioned, hindering us to achieve good image quality. In this paper, we propose a novel Bayesian method to tackle the bottleneck in CB-XLCT reconstruction. The method utilizes a local regularization strategy based on Gaussian Markov random field to mitigate the ill-conditioness of CB-XLCT. An alternating optimization scheme is then used to automatically calculate all the unknown hyperparameters while an iterative coordinate descent algorithm is adopted to reconstruct the image with a voxel-based closed-form solution. Results of numerical simulations and mouse experiments show that the self-adaptive Bayesian method significantly improves the CB-XLCT image quality as compared with conventional methods.

  20. Computed tomography for light materials using a monochromatic X-ray beam produced by parametric X-ray radiation

    Energy Technology Data Exchange (ETDEWEB)

    Hayakawa, Y., E-mail: yahayak@lebra.nihon-u.ac.jp [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Hayakawa, K.; Inagaki, M. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Kaneda, T. [Nihon University School of Dentistry at Matsudo, Sakaecho-Nishi 2-870-1, Matsudo 271-8587 (Japan); Nakao, K.; Nogami, K. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Sakae, T. [Nihon University School of Dentistry at Matsudo, Sakaecho-Nishi 2-870-1, Matsudo 271-8587 (Japan); Sakai, T.; Sato, I. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Takahashi, Y. [Institute of Materials Structure Science, High Energy Accelerator Research Organization (KEK), Oho, Tsukuba 305-8501 (Japan); Tanaka, T. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan)

    2013-08-15

    Computed tomography (CT) for light materials such as soft biological tissues was performed using a monochromatic X-ray beam provided by a parametric X-ray radiation (PXR) source at the Laboratory for Electron Beam Research and Application (LEBRA) of Nihon University. Using a high-efficiency flat panel detector (FPD), each projection image for CT was taken with exposure times of 5 or 10 s, and 60–360 projection images in each run were obtained with total measurement time of 5 min to 1 h. CT images were obtained from the projection images using the conventional calculation method. The typical tomograms obtained had sharp outlines, which are likely attributable to the propagation-based phase contrast.

  1. Design and implementation of an x-ray strain measurement capability using a rotating anode machine

    Energy Technology Data Exchange (ETDEWEB)

    Roberts, J.A.; Rangaswamy, P.; Lujan, M. Jr.; Bourke, M.A.M.

    1998-12-31

    This is the final report of a three-year, Laboratory Directed Research and Development (LDRD) project at the Los Alamos National Laboratory (LANL). Residual stresses close to the surface can improve the reliability and lifetime of parts for technological applications. X-ray diffraction plays a significant role in gaining an exact knowledge of the stresses at the surface and their depth distribution. An x-ray capability at Los Alamos is key to developing and maintaining industrial collaborations in strain effects. To achieve this goal, the authors implemented a residual strain measuring station on the rotating anode x-ray instrument at the Lujan Center. This capability has been used to investigate residual strains in heat treated automotive components, machining effects on titanium alloys, resistance welded steel joints, titanium matrix fiber reinforced composites, ceramic matrix composites, thin films, and ceramic coatings. The overall objective is to combine both x-ray and neutron diffraction measurements with numerical models (e.g., finite element calculations).

  2. Ion beam induced surface graphitization of CVD diamond for x-ray beam position monitor applications

    International Nuclear Information System (INIS)

    Liu, Chian; Shu, D.; Kuzay, T.M.; Wen, L.; Melendres, C.A.; Argonne National Lab., IL

    1996-01-01

    The Advanced Photon Source at ANL is a third-generation synchrotron facility that generates powerful x-ray beams on its undulator beamlines. It is important to know the position and angle of the x- ray beam during experiments. Due to very high heat flux levels, several patented x-ray beam position monitors (XBPM) exploiting chemical vapor deposition (CVD) diamond have been developed. These XBPMs have a thin layer of low-atomic-mass metallic coating so that photoemission from the x rays generate a minute but measurable current for position determination. Graphitization of the CVD diamond surface creates a very thin, intrinsic and conducting layer that can stand much higher temperatures and minimal x-ray transmission losses compared to the coated metallic layers. In this paper, a laboratory sputter ion source was used to transform selected surfaces of a CVD diamond substrate into graphite. The effect of 1-5 keV argon ion bombardment on CVD diamond surfaces at various target temperatures from 200 to 500 C was studied using Auger electron spectroscopy and in-situ electrical resistivity measurements. Graphitization after the ion bombardment has been confirmed and optimum conditions for graphitization studied. Raman spectroscopy was used to identify the overall diamond structure in the bulk of CVD diamond substrate after the ion bombardments. It was found that target temperature plays an important role in stability and electrical conductivity of the irradiated CVD diamonds

  3. Generation of low KV x-ray portal images with mega-voltage electron beams

    International Nuclear Information System (INIS)

    Kenny, J.; Ebert, M.

    2004-01-01

    phantom and a purpose built contrast phantom were evaluated for contrast enhancement using purpose written MATLAB code. Information from these images was implemented in the Monte Carlo model to refine the target construction. Electron mode image acquisition was achieved using the clinical (R and V) and service interfaces however the EPID was activated using an internal trigger rather then a trigger pulse from the linac, since it does not recognise the linac pulses for electron modes. Images could be acquired in as little as 5 monitor units (MU), with reasonable quality images generated with ∼20MU. Doses of this magnitude indicate that such images could be acquired in a clinically viable amount of time, especially if higher dose rates was used e.g. 1000MU/min. Image quality is affected because the EPID is not synchronised to the linac pulses. Furthermore a flood-field correction image for the low energy beam is difficult to acquire and a 'standard' one has been applied. Although not optimal for efficient x-ray production, the low Z target maximises the content of low energy x-rays in the imaging beam, leading to portal images with improved contrast. Using a high efficiency EPID, images could be acquired in clinically practical times. This method of low energy portal imaging is relatively simple, inexpensive to construct and easy to install. Image quality would benefit from a special imaging mode that maximizes EPID performance under these conditions. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine

  4. X-ray convergent beam pattern simulation using the Moodie-Wagenfeld equations: 3-beam Laue case

    International Nuclear Information System (INIS)

    Liu, L.; Goodman, P.

    1998-01-01

    Pattern simulations for 3-beam X-ray diffraction are presented, by multi-slice calculations based on Moodie and Wagenfeld's formulation of the X-ray equations, which factorise Maxwell's equations into Dirac format, using circular-polarisation bases. The results are presented in the form of convergent-beam patterns for each diffraction order, using experience gained from CBED (convergent beam electron diffraction) and LACBED (large-angle CBED), since this displays the results in the most compact form. The acronym CBXRAD (convergent-beam X-ray-diffraction) is used for these patterns. Although optics required for the complete patterns is not currently available, capillary focussing is undergoing rapid development, and our simulations define critical angular ranges within reach of current designs. Simulations for light and heavy-atoms structures belonging to the enantiomorphic space-group pair P3 1 21 and P3 2 21, provide clear evidence of chiral interaction between radiation and structure, highlighting divergences from the well studied CBED pattern symmetries. MoKα 1 and TaKα 1 wavelengths were used to minimise absorption for the two structures respectively, although 'anomalous absorption' is always important due to the large thicknesses required (up to 20 mm)

  5. Diamond monochromator for high heat flux synchrotron x-ray beams

    International Nuclear Information System (INIS)

    Khounsary, A.M.; Smither, R.K.; Davey, S.; Purohit, A.

    1992-12-01

    Single crystal silicon has been the material of choice for x-ray monochromators for the past several decades. However, the need for suitable monochromators to handle the high heat load of the next generation synchrotron x-ray beams on the one hand and the rapid and on-going advances in synthetic diamond technology on the other make a compelling case for the consideration of a diamond mollochromator system. In this Paper, we consider various aspects, advantage and disadvantages, and promises and pitfalls of such a system and evaluate the comparative an monochromator subjected to the high heat load of the most powerful x-ray beam that will become available in the next few years. The results of experiments performed to evaluate the diffraction properties of a currently available synthetic single crystal diamond are also presented. Fabrication of diamond-based monochromator is within present technical means

  6. Flat Field Anomalies in an X-ray CCD Camera Measured Using a Manson X-ray Source (HTPD 08 paper)

    International Nuclear Information System (INIS)

    Haugh, M; Schneider, M B

    2008-01-01

    The Static X-ray Imager (SXI) is a diagnostic used at the National Ignition Facility (NIF) to measure the position of the X-rays produced by lasers hitting a gold foil target. The intensity distribution taken by the SXI camera during a NIF shot is used to determine how accurately NIF can aim laser beams. This is critical to proper NIF operation. Imagers are located at the top and the bottom of the NIF target chamber. The CCD chip is an X-ray sensitive silicon sensor, with a large format array (2k x 2k), 24 (micro)m square pixels, and 15 (micro)m thick. A multi-anode Manson X-ray source, operating up to 10kV and 10W, was used to characterize and calibrate the imagers. The output beam is heavily filtered to narrow the spectral beam width, giving a typical resolution E/ΔE ∼ 10. The X-ray beam intensity was measured using an absolute photodiode that has accuracy better than 1% up to the Si K edge and better than 5% at higher energies. The X-ray beam provides full CCD illumination and is flat, within ±1% maximum to minimum. The spectral efficiency was measured at 10 energy bands ranging from 930 eV to 8470 eV. We observed an energy dependent pixel sensitivity variation that showed continuous change over a large portion of the CCD. The maximum sensitivity variation occurred at 8470 eV. The geometric pattern did not change at lower energies, but the maximum contrast decreased and was not observable below 4 keV. We were also able to observe debris, damage, and surface defects on the CCD chip. The Manson source is a powerful tool for characterizing the imaging errors of an X-ray CCD imager. These errors are quite different from those found in a visible CCD imager

  7. The survey of the surface doses of the dental x-ray machines

    International Nuclear Information System (INIS)

    Lee, Jae Seo; Kang, Byung Cheol; Yoon, Suk Ja

    2005-01-01

    The purpose of this study was to investigate variability of doses with same exposure parameters and evaluate radiographic density according to the variability of doses. Twenty-eight MAX-GLS (Shinhung Co, Seoul, Korea), twenty-one D-60-S (DongSeo Med, Seoul, Korea), and eleven REX-601 (Yoshida Dental MFG, Tokyo, Japan) dental x-ray machines were selected for this study. Surface doses were measured under selected combinations of tube voltage, tube current, exposure time, and constant distance 42 cm from the focal spot to the surface of the Multi-O-meter (Unfors Instrument, Billdal, Sweden). Radiographic densities were measured on the films at maximum, minimum and mean surface doses of each brand of x-ray units. With MAX-GLS, the maximum surface doses were thirteen to fourteen times as much as the minimum surfaces doses. With D-60-S, the maximum surface doses were three to eight times as much as the minimum surface doses. With REX-601, the maximum surface doses were six to ten times as much as the minimum surface doses. The differences in radiographic densities among maximum, mean, and minimum doses were significant (p<0.01). The surface exposure doses of each x-ray machine at the same exposure parameters were different within the same manufacturer's machines.

  8. On the response of electronic personal dosimeters in constant potential and pulsed X-ray beams

    Energy Technology Data Exchange (ETDEWEB)

    Guimaraes, Margarete C.; Silva, Teogenes; Silva, Claudete R.E., E-mail: margaretecristinag@gmail.com [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil); Oliveira, Paulo Marcio C. de [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Anatomia e Imagem

    2015-07-01

    Electronic personal dosimeters (EPDs) based on solid state detectors have widely been used but some deficiencies in their response in pulsed radiation beams have been reported. Nowadays, there is not an international standard for pulsed X-ray beams for calibration or type testing of dosimeters. Irradiation conditions for testing the response of EPDs in both the constant potential and pulsed X-ray beams were established in CDTN. Three different types of EPDs were tested in different conditions in similar ISO and IEC X-ray qualities. Results stressed the need of performing additional checks before using EPDs in constant potential or pulsed X-rays. (author)

  9. Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography

    International Nuclear Information System (INIS)

    Speidel, Michael A.; Wilfley, Brian P.; Star-Lack, Josh M.; Heanue, Joseph A.; Van Lysel, Michael S.

    2006-01-01

    The scanning-beam digital x-ray (SBDX) system is designed for x-ray dose reduction in cardiac angiographic applications. Scatter reduction, efficient detection of primary x-rays, and an inverse beam geometry are the main components of the entrance dose reduction strategy. This paper reports the construction of an SBDX prototype, image reconstruction techniques, and measurements of spatial resolution and x-ray output. The x-ray source has a focal spot that is electronically scanned across a large-area transmission target. A multihole collimator beyond the target defines a series of x-ray beams directed at a distant small-area detector array. The prototype has a 23 cmx23 cm target, 100x100 focal spot positions, and a 5 cmx5 cm CdTe detector positioned 150 cm from the target. With this nonmechanical method of beam scanning, patient images with low detected scatter are generated at up to 30 frame/s. SBDX data acquisition is tomosynthetic. The prototype simultaneously reconstructs 16 planes spaced throughout the cardiac volume using shift-and-add backprojection. Image frames analogous to conventional projection images are generated with a multiplane compositing algorithm. Single-plane versus multiplane reconstruction of contrast-filled coronary arteries is demonstrated with images of the porcine heart. Phantom and porcine imaging studies show multiplane reconstruction is practicable under clinically realistic levels of patient attenuation and cardiac motion. The modulation transfer function for an in-plane slit at mechanical isocenter measured 0.41-0.56 at 1 cycle/mm, depending on the detector element to image pixel interpolation technique. Modeling indicates that desired gains in spatial resolution are achievable by halving the detector element width. The x-ray exposure rate 15 cm below isocenter, without table or patient in the beam, measured 11.5 R/min at 120 kVp, 24.3 kWp and 3.42 R/min at 70 kVp, 14.2 kWp

  10. Scattered hard X-ray and γ-ray generation from a chromatic electron beam

    Energy Technology Data Exchange (ETDEWEB)

    Coleman, J. E. [Los Alamos National Laboratory, Los Alamos, New Mexico 87545 (United States); Welch, D. R.; Miller, C. L. [Voss Scientific, Albuquerque, New Mexico 87108 (United States)

    2015-11-14

    An array of photon diagnostics has been deployed on a high power relativistic electron beam diode. Electrons are extracted through a 17.8 cm diode from the surface discharge of a carbon fiber velvet cathode with a nominal diode voltage of 3.8 MV. <10% of the 100 ns electron pulse is composed of off energy electrons (1–3 MeV) accelerated during the rise and fall of the pulse that impact the stainless steel beam pipe and generate a Bremsstrahlung spectrum of 0.1–3 MeV photons with a total count of 10{sup 11}. The principal objective of these experiments is to quantify the electron beam dynamics and spatial dynamics of the hard X-ray and γ-ray flux generated in the diode region. A qualitative comparison of experimental and calculated results are presented, including time and energy resolved electron beam propagation and scattered photon measurements with X-ray PIN diodes and a photomultiplier tube indicating a dose dependence on the diode voltage >V{sup 4} and detected photon counts of nearly 10{sup 6} at a radial distance of 1 m which corresponds to dose ∼40 μrad at 1 m.

  11. Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016

    Science.gov (United States)

    Bachtiar, Ilham; Abdullah, Bualkar; Tahir, Dahlan

    2018-03-01

    This paper describes the conventional X-ray machine parameters tested in the region of South Sulawesi from 2014 to 2016. The objective of this research is to know deviation of every parameter of conventional X-ray machine. The testing parameters were analyzed by using quantitative methods with participatory observational approach. Data collection was performed by testing the output of conventional X-ray plane using non-invasive x-ray multimeter. The test parameters include tube voltage (kV) accuracy, radiation output linearity, reproducibility and radiation beam value (HVL) quality. The results of the analysis show four conventional X-ray test parameters have varying deviation spans, where the tube voltage (kV) accuracy has an average value of 4.12%, the average radiation output linearity is 4.47% of the average reproducibility of 0.62% and the averaged of the radiation beam (HVL) is 3.00 mm.

  12. Broad beam X-rays attenuation in silicum glass

    International Nuclear Information System (INIS)

    Risticj, Dj.; Vukovicj, S.; Markovicj, P.

    1987-01-01

    Using broad beam geometry the attenuation for domestic silicum glass have been studied for constant X-ray potentials from 50 to 150 kV. The density of the silicium glass was 2,5x10 3 kg/m 3 . From the attenuation curves the half value layers were obtained. The use of this glass as the biological shield is pointed out. (author). 2 refs.; 2 tabs.; 2 figs

  13. Beam hardening: Analytical considerations of the effective attenuation coefficient of x-ray tomography

    International Nuclear Information System (INIS)

    Alles, J.; Mudde, R. F.

    2007-01-01

    Polychromatic x-ray beams traveling though material are prone to beam hardening, i.e., the high energy part of the incident spectrum gets over represented when traveling farther into the material. This study discusses the concept of a mean attenuation coefficient in a formal way. The total energy fluence is one-to-one related to the traveled distance in case of a polychromatic beam moving through a given, inhomogeneous material. On the basis of this one-to-one relation, it is useful to define a mean attenuation coefficient and study its decrease with depth. Our results are based on a novel parametrization of the energy dependence of the attenuation coefficient that allows for closed form evaluation of certain spectral integrals. This approach underpins the ad hoc semianalytical expressions given in the literature. An analytical model for the average attenuation coefficient is proposed that uses a simple fit of the attenuation coefficient as a function of the photon energy as input. It is shown that a simple extension of this model gives a rather good description of beam hardening for x-rays traveling through water

  14. Impact of large x-ray beam collimation on image quality

    Science.gov (United States)

    Racine, Damien; Ba, Alexandre; Ott, Julien G.; Bochud, François O.; Verdun, Francis R.

    2016-03-01

    Large X-ray beam collimation in computed tomography (CT) opens the way to new image acquisition techniques and improves patient management for several clinical indications. The systems that offer large X-ray beam collimation enable, in particular, a whole region of interest to be investigated with an excellent temporal resolution. However, one of the potential drawbacks of this option might be a noticeable difference in image quality along the z-axis when compared with the standard helical acquisition mode using more restricted X-ray beam collimations. The aim of this project is to investigate the impact of the use of large X-ray beam collimation and new iterative reconstruction on noise properties, spatial resolution and low contrast detectability (LCD). An anthropomorphic phantom and a custom made phantom were scanned on a GE Revolution CT. The images were reconstructed respectively with ASIR-V at 0% and 50%. Noise power spectra, to evaluate the noise properties, and Target Transfer Functions, to evaluate the spatial resolution, were computed. Then, a Channelized Hotelling Observer with Gabor and Dense Difference of Gaussian channels was used to evaluate the LCD using the Percentage correct as a figure of merit. Noticeable differences of 3D noise power spectra and MTF have been recorded; however no significant difference appeared when dealing with the LCD criteria. As expected the use of iterative reconstruction, for a given CTDIvol level, allowed a significant gain in LCD in comparison to ASIR-V 0%. In addition, the outcomes of the NPS and TTF metrics led to results that would contradict the outcomes of CHO model observers if used for a NPWE model observer (Non- Prewhitening With Eye filter). The unit investigated provides major advantages for cardiac diagnosis without impairing the image quality level of standard chest or abdominal acquisitions.

  15. PBFA II lithium beam characterization from inner-shell x-ray images

    International Nuclear Information System (INIS)

    Moats, A.R.; Derzon, M.S.; Chandler, G.A.; Dukart, R.J.; Haill, T.A.

    1994-01-01

    The Particle Beam Fusion Accelerator (PBFA II) is not driving targets with ICF-relevant lithium ion beams. During the most recent lithium beam target series, time-integrated x-ray pinhole cameras viewed the ion-induced inner-shell x-ray fluorescence from the central gold cone target and a titanium-coated strip. Ion beam profiles at a nominal 10 mm radius and fixed azimuthal direction were obtained from images of the Ti K α , fluorescence of a Ti-coated Al diagnostic wire. The gold cone gave us beam profiles at a nominal 3 mm radius and at all azimuthal angles from the Au L α fluorescence. From these profiles, we obtained the ion beam vertical focus position, full-width-at-half-maximum, and the degree of azimuthal uniformity for the lithium target shots. For these initial results, beam steering problems were evident. Azimuthal uniformity was measured from the ion beam footprint on the outer Au case (predominantly Au L α ) of the hohlraum target and were found to be in the same range (up to 30%) as for previous proton beam target series. We then present plans for Li beam diagnostics for an upcoming target experimental series

  16. X-ray beam monitor made by thin-film CVD single-crystal diamond.

    Science.gov (United States)

    Marinelli, Marco; Milani, E; Prestopino, G; Verona, C; Verona-Rinati, G; Angelone, M; Pillon, M; Kachkanov, V; Tartoni, N; Benetti, M; Cannatà, D; Di Pietrantonio, F

    2012-11-01

    A novel beam position monitor, operated at zero bias voltage, based on high-quality chemical-vapor-deposition single-crystal Schottky diamond for use under intense synchrotron X-ray beams was fabricated and tested. The total thickness of the diamond thin-film beam monitor is about 60 µm. The diamond beam monitor was inserted in the B16 beamline of the Diamond Light Source synchrotron in Harwell (UK). The device was characterized under monochromatic high-flux X-ray beams from 6 to 20 keV and a micro-focused 10 keV beam with a spot size of approximately 2 µm × 3 µm square. Time response, linearity and position sensitivity were investigated. Device response uniformity was measured by a raster scan of the diamond surface with the micro-focused beam. Transmissivity and spectral responsivity versus beam energy were also measured, showing excellent performance of the new thin-film single-crystal diamond beam monitor.

  17. Adjustment and Prediction of X-Ray Machine Factors Based on Neural Artificial Inculcating

    International Nuclear Information System (INIS)

    Hussein, A.Z.; Amin, E.S.; Ibrahim, M.S.

    2009-01-01

    Since the discovery of X-rays, their use in examination has become an integral part of medical diagnostic radiology. The use of X-rays is harmful to human beings but recent technological advances and regulatory constraints have made the medical X-rays much safer than they were at the beginning of the 20th century. However, the potential benefits of the engineered safety features can not be fully realized unless the operators are aware of these safety features. The aim of this work is to adjust and predict X-ray machine factors (current and voltage) using neural artificial network in order to obtain effective dose within the range of dose limitation system and assure radiological safety.

  18. An extended range soft X-ray beam line for the 1 GeV storage ring Aladdin

    International Nuclear Information System (INIS)

    Hulbert, S.L.; Stott, J.P.; Brown, F.C.

    1983-01-01

    The design and implementation of a soft X-ray beam line on the new 1 GeV storage ring Aladdin in Stoughton, Wisconsin is discussed. The beam line consists of a long horizontally focussing collection mirror, an extended range (50-1500 eV) grasshopper monochromator, an ellipsoidal refocussing mirror, and a photoemission chamber. Also discussed are the factors considered in matching the monochromator to the storage ring, flux and performance expectations, and the results of a ray tracing analysis. (orig.)

  19. Optical and x-ray imaging of electron beams using synchrotron emission

    International Nuclear Information System (INIS)

    Wilke, M.

    1995-01-01

    In the case of very low emittance electron and positron storage ring beams, it is impossible to make intrusive measurements of beam properties without increasing the emittance and possibly disrupting the beam. In cases where electron or positron beams have high average power densities (such as free electron laser linacs), intrusive probes such as wires and optical transition radiation screens or Cherenkov emitting screens can be easily damaged or destroyed. The optical and x-ray emissions from the bends in the storage rings and often from linac bending magnets can be used to image the beam profile to obtain emittance information about the beam. The techniques, advantages and limitations of using both optical and x-ray synchrotron emission to measure beam properties are discussed and the possibility of single bunch imaging is considered. The properties of suitable imagers and converters such as phosphors are described. Examples of previous, existing and planned applications are given where available, including a pinhole imaging system currently being designed for the Advanced Photon Source at Argonne National Laboratory

  20. Optical and x-ray imaging of electron beams using synchrotron emission

    International Nuclear Information System (INIS)

    Wilke, M.D.

    1994-01-01

    In the case of very low eniittance electron and positron storage ring beams, it is impossible to make intrusive measurements of beam properties without increasing the emittance and possibly disrupting the beam. In cases where electron or positron beams have high average power densities (such as free electron laser linacs), intrusive probes such as wires and optical transition radiation screens or Cherenkov emitting screens can be easily damaged or destroyed. The optical and x-ray emissions from the bends in the storage rings and often from linac bending magnets can be used to image the beam profile to obtain emittance information about the beam. The techniques, advantages and limitations of using both optical and x-ray synchrotron emission to measure beam properties are discussed and the possibility of single bunch imaging is considered. The properties of suitable imagers and converters such as phosphors are described. Examples of previous, existing and planned applications are given where available, including a pinhole imaging system currently being designed for the Advanced Photon Source at Argonne National Laboratory

  1. Pin-photodiode array for the measurement of fan-beam energy and air kerma distributions of X-ray CT scanners.

    Science.gov (United States)

    Haba, Tomonobu; Koyama, Shuji; Aoyama, Takahiko; Kinomura, Yutaka; Ida, Yoshihiro; Kobayashi, Masanao; Kameyama, Hiroshi; Tsutsumi, Yoshinori

    2016-07-01

    Patient dose estimation in X-ray computed tomography (CT) is generally performed by Monte Carlo simulation of photon interactions within anthropomorphic or cylindrical phantoms. An accurate Monte Carlo simulation requires an understanding of the effects of the bow-tie filter equipped in a CT scanner, i.e. the change of X-ray energy and air kerma along the fan-beam arc of the CT scanner. To measure the effective energy and air kerma distributions, we devised a pin-photodiode array utilizing eight channels of X-ray sensors arranged at regular intervals along the fan-beam arc of the CT scanner. Each X-ray sensor consisted of two plate type of pin silicon photodiodes in tandem - front and rear photodiodes - and of a lead collimator, which only allowed X-rays to impinge vertically to the silicon surface of the photodiodes. The effective energy of the X-rays was calculated from the ratio of the output voltages of the photodiodes and the dose was calculated from the output voltage of the front photodiode using the energy and dose calibration curves respectively. The pin-photodiode array allowed the calculation of X-ray effective energies and relative doses, at eight points simultaneously along the fan-beam arc of a CT scanner during a single rotation of the scanner. The fan-beam energy and air kerma distributions of CT scanners can be effectively measured using this pin-photodiode array. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  2. Spectral structure of a polycapillary lens shaped X-ray beam

    Science.gov (United States)

    Gogolev, A. S.; Filatov, N. A.; Uglov, S. R.; Hampai, D.; Dabagov, S. B.

    2018-04-01

    Polycapillary X-ray optics is widely used in X-ray analysis techniques to create a small secondary source, for instance, or to deliver X-rays to the point of interest with minimum intensity losses [1]. The main characteristics of the analytical devices on its base are the size and divergence of the focused or translated beam. In this work, we used the photon-counting pixel detector ModuPIX to study the parameters for polycapillary focused X-ray tube radiation as well as the energy and spatial dependences of radiation at the focus. We have characterized the high-speed spectral camera ModuPIX, which is a single Timepix device with a fast parallel readout allowing up to 850 frames per second with 256 × 256 pixels and a 55 μm pitch defined by the frame frequency. By means of the silicon monochromator the energy response function is measured in clustering mode by the energy scan over total X-ray tube spectrum.

  3. Improving x-ray fluorescence signal for benchtop polychromatic cone-beam x-ray fluorescence computed tomography by incident x-ray spectrum optimization: a Monte Carlo study.

    Science.gov (United States)

    Manohar, Nivedh; Jones, Bernard L; Cho, Sang Hyun

    2014-10-01

    To develop an accurate and comprehensive Monte Carlo (MC) model of an experimental benchtop polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) setup and apply this MC model to optimize incident x-ray spectrum for improving production/detection of x-ray fluorescence photons from gold nanoparticles (GNPs). A detailed MC model, based on an experimental XFCT system, was created using the Monte Carlo N-Particle (MCNP) transport code. The model was validated by comparing MC results including x-ray fluorescence (XRF) and scatter photon spectra with measured data obtained under identical conditions using 105 kVp cone-beam x-rays filtered by either 1 mm of lead (Pb) or 0.9 mm of tin (Sn). After validation, the model was used to investigate the effects of additional filtration of the incident beam with Pb and Sn. Supplementary incident x-ray spectra, representing heavier filtration (Pb: 2 and 3 mm; Sn: 1, 2, and 3 mm) were computationally generated and used with the model to obtain XRF/scatter spectra. Quasimonochromatic incident x-ray spectra (81, 85, 90, 95, and 100 keV with 10 keV full width at half maximum) were also investigated to determine the ideal energy for distinguishing gold XRF signal from the scatter background. Fluorescence signal-to-dose ratio (FSDR) and fluorescence-normalized scan time (FNST) were used as metrics to assess results. Calculated XRF/scatter spectra for 1-mm Pb and 0.9-mm Sn filters matched (r ≥ 0.996) experimental measurements. Calculated spectra representing additional filtration for both filter materials showed that the spectral hardening improved the FSDR at the expense of requiring a much longer FNST. In general, using Sn instead of Pb, at a given filter thickness, allowed an increase of up to 20% in FSDR, more prominent gold XRF peaks, and up to an order of magnitude decrease in FNST. Simulations using quasimonochromatic spectra suggested that increasing source x-ray energy, in the investigated range of 81-100 ke

  4. Electron beam produced in a transient hollow cathode discharge: beam electron distribution function, X-ray emission and solid target ablation

    International Nuclear Information System (INIS)

    Nistor, Magdalena

    2000-01-01

    This research thesis aims at a better knowledge of phenomena occurring during transient hollow cathode discharges. The author first recalls the characteristics of such a discharge which make it different from conventional pseudo-spark discharges. The objective is to characterise the electron beam produced within the discharge, and the phenomena associated with its interaction with a solid or gaseous target, leading to the production of an X ray or visible radiation. Thus, the author reports the measurement (by magnetic deflection) of the whole time-averaged electronic distribution function. Such a knowledge is essential for a better use of the electron beam in applications such as X-ray source or material ablation. As high repetition frequency pulse X ray sources are very interesting tools, he reports the development and characterisation of Bremsstrahlung X rays during a beam-target interaction. He finally addresses the implementation of a spectroscopic diagnosis for the filamentary plasma and the ablation of a solid target by the beam [fr

  5. In situ micro-focused X-ray beam characterization with a lensless camera using a hybrid pixel detector

    International Nuclear Information System (INIS)

    Kachatkou, Anton; Marchal, Julien; Silfhout, Roelof van

    2014-01-01

    Position and size measurements of a micro-focused X-ray beam, using an X-ray beam imaging device based on a lensless camera that collects radiation scattered from a thin foil placed in the path of the beam at an oblique angle, are reported. Results of studies on micro-focused X-ray beam diagnostics using an X-ray beam imaging (XBI) instrument based on the idea of recording radiation scattered from a thin foil of a low-Z material with a lensless camera are reported. The XBI instrument captures magnified images of the scattering region within the foil as illuminated by the incident beam. These images contain information about beam size, beam position and beam intensity that is extracted during dedicated signal processing steps. In this work the use of the device with beams for which the beam size is significantly smaller than that of a single detector pixel is explored. The performance of the XBI device equipped with a state-of-the-art hybrid pixel X-ray imaging sensor is analysed. Compared with traditional methods such as slit edge or wire scanners, the XBI micro-focused beam characterization is significantly faster and does not interfere with on-going experiments. The challenges associated with measuring micrometre-sized beams are described and ways of optimizing the resolution of beam position and size measurements of the XBI instrument are discussed

  6. Generation and focusing of high energy, 35-kA electron beams for pulsed-diode radiographic machines: Theory and experiment

    International Nuclear Information System (INIS)

    Carlson, R.L.; George, M.J.; Hughes, T.P.; Welch, D.R.

    1993-01-01

    Cathode ball and anode planar-foil geometries used to generate self-focused beams onto x-ray conversion targets via beam-induced ionization in gas cells have been investigated since the early 1970's by J. C. Martin et al at Aldermaston, U.K. The building of a succession of increasingly higher voltage, pulsed-diode machines tailored for flash x radiography has resulted. Given sufficient dose to penetrate an object, the spot size of the x-ray source generally determines the resolution of a radiograph. Reported are particle-in-cell code simulations applied to beam generation in the A-K gap and the self-focusing onto the target. A Monte Carlo code for neutron, photon, and electron transport converts the beam particles at the target to photons with transport to a film plane used to calculate the spot size. Comparisons are made to experiments using the Ector (3.5--4 MeV) and PIXY (4--8 MeV) pulsed-diode radiographic machines at Los Alamos

  7. An investigation of dose changes for therapeutic kilovoltage x-ray beams with underlying lead shielding

    International Nuclear Information System (INIS)

    Hill, Robin; Healy, Brendan; Holloway, Lois; Baldock, Clive

    2007-01-01

    Kilovoltage x-ray beams are used to treat cancer on or close to the skin surface. Many clinical cases use high atomic number materials as shielding to reduce dose to underlying healthy tissues. In this work, we have investigated the effect on both the surface dose and depth doses in a water phantom with lead shielding at depth in the phantom. The EGSnrc Monte Carlo code was used to simulate the water phantom and to calculate the surface doses and depth doses using primary x-ray beam spectra derived from an analytical model. The x-ray beams were in the energy range of 75-135 kVp with field sizes of 2, 5 and 8 cm diameter. The lead sheet was located beneath the water surface at depths ranging from 0.5-7.5 cm. The surface dose decreased as the lead was positioned closer to the water surface and as the field size was increased. The variation in surface dose as a function of x-ray beam energy was only small but the maximum reduction occurred for the 100 kVp x-ray beam. For the 8 cm diameter field with the lead at 1 cm depth and using the 100 kVp x-ray beam, the surface dose was reduced to 0.898 of the surface dose in the water phantom only. Measured surface dose changes, using a Farmer-type ionization chamber, agreed with the Monte Carlo calculated doses. Calculated depth doses in water with a lead sheet positioned below the surface showed that the dose fall-off increased as the lead was positioned closer to the water surface as compared to the depth dose in the water phantom only. Monte Carlo calculations of the total x-ray beam spectrum at the water surface showed that the total fluence decreased due to a reduction in backscatter from within the water and very little backscatter from the lead. The mean energy of the x-ray spectrum varied less than 1 keV, with the lead at 1 cm beneath the water phantom surface. As the Monte Carlo calculations showed good agreement with the measured results, this method can be used to verify surface dose changes in clinical situations

  8. Fabrication techniques of X-ray spiral zone plates

    International Nuclear Information System (INIS)

    Gao Nan; Zhu Xiaoli; Li Hailiang; Xie Changqing

    2010-01-01

    The techniques to make X-ray spiral zone plates using electron beam and X-ray lithography were studied. A master mask was fabricated on polyimide membrane by E-beam lithography and micro-electroplating. Spiral zone plates were efficiently replicated by X-ray lithography and micro-electroplating. By combining the techniques, spiral zone plates at 1 keV were successfully fabricate. With an outermost zone width of the 200 nm, and the gold absorbers thickness of 700 nm, the high quality zone plates can be used for X-ray phase contrast microscopy.(authors)

  9. Beam dynamics simulation in the X-ray Compton source

    International Nuclear Information System (INIS)

    Gladkikh, P.; Karnaukhov, I.; Telegin, Yu.; Shcherbakov, A.; Zelinsky, A.

    2002-01-01

    At the National Science Center 'Kharkov Institute of Physics and Technology' the X-ray source based on Compton scattering has been developed. The computer code for simulation of electron beam dynamics with taking into account the Compton scattering effect based on Monte Carlo method is described in this report. The first results of computer simulation of beam dynamics with electron-photon interaction, parameters of electron and photon beams are presented. Calculations were carried out with the lattice of synchrotron light source SRS-800 Ukrainian Synchrotron Center

  10. Beam dynamics simulation in the X-ray Compton source

    CERN Document Server

    Gladkikh, P; Telegin, Yu P; Shcherbakov, A; Zelinsky, A

    2002-01-01

    At the National Science Center 'Kharkov Institute of Physics and Technology' the X-ray source based on Compton scattering has been developed. The computer code for simulation of electron beam dynamics with taking into account the Compton scattering effect based on Monte Carlo method is described in this report. The first results of computer simulation of beam dynamics with electron-photon interaction, parameters of electron and photon beams are presented. Calculations were carried out with the lattice of synchrotron light source SRS-800 Ukrainian Synchrotron Center.

  11. Bremsstrahlung X rays from Jovian auroral electrons

    International Nuclear Information System (INIS)

    Barbosa, D.D.

    1990-01-01

    The spectrum of X rays from the planet Jupiter is calculated according to an auroral electron beam model. The electrons are assumed to be accelerated by a field-aligned potential drop and penetrate into the atmosphere as a Maxwellian beam of primaries which are scattered, degraded in energy, and merged with a population of ionization secondaries having a power law energy distribution. The soft X rays observed by the Einstein Observatory satelltie are due to bremsstrahlung from the sedondary electrons in the H 2 atmosphere. A good match to the X ray data is obtained if the power law spectral index of the secondary electrons, γ e , is ≅ 2, yielding a power law slope for the photon flux γ X = γ e + 1 ≅ 3. The X ray intensity is best reconciled with a beam of primaries having a characteristic energy 30-100 keV and penetrating the homopause with an auroral energy flux typically of 10-20 ergs cm -2 s -1 but no greater than 50 ergs cm -2 s -1

  12. Multi-mounted X-ray cone-beam computed tomography

    Science.gov (United States)

    Fu, Jian; Wang, Jingzheng; Guo, Wei; Peng, Peng

    2018-04-01

    As a powerful nondestructive inspection technique, X-ray computed tomography (X-CT) has been widely applied to clinical diagnosis, industrial production and cutting-edge research. Imaging efficiency is currently one of the major obstacles for the applications of X-CT. In this paper, a multi-mounted three dimensional cone-beam X-CT (MM-CBCT) method is reported. It consists of a novel multi-mounted cone-beam scanning geometry and the corresponding three dimensional statistical iterative reconstruction algorithm. The scanning geometry is the most iconic design and significantly different from the current CBCT systems. Permitting the cone-beam scanning of multiple objects simultaneously, the proposed approach has the potential to achieve an imaging efficiency orders of magnitude greater than the conventional methods. Although multiple objects can be also bundled together and scanned simultaneously by the conventional CBCT methods, it will lead to the increased penetration thickness and signal crosstalk. In contrast, MM-CBCT avoids substantially these problems. This work comprises a numerical study of the method and its experimental verification using a dataset measured with a developed MM-CBCT prototype system. This technique will provide a possible solution for the CT inspection in a large scale.

  13. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    Energy Technology Data Exchange (ETDEWEB)

    Wilson, Matthew D., E-mail: Matt.Wilson@stfc.ac.uk; Seller, Paul; Veale, Matthew C. [Science and Technology Facilities Council, Rutherford Appleton Laboratory, Harwell Campus,UK (United Kingdom); Connolley, Thomas [Diamond Light Source, I12 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal [Diamond Light Source, B16 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Grant, Patrick S.; Liotti, Enzo; Lui, Andrew [Department of Materials, University of Oxford Parks Road, Oxford (United Kingdom)

    2016-07-27

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm{sup 2} with one of the 80×80 pixels imaging an area equivalent to 13µm{sup 2}. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  14. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    International Nuclear Information System (INIS)

    Wilson, Matthew D.; Seller, Paul; Veale, Matthew C.; Connolley, Thomas; Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal; Grant, Patrick S.; Liotti, Enzo; Lui, Andrew

    2016-01-01

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm"2 with one of the 80×80 pixels imaging an area equivalent to 13µm"2. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  15. Experimental characterization of X-ray transverse coherence in the presence of beam transport optics

    DEFF Research Database (Denmark)

    Chubar, O.; Fluerasu, A.; Chu, Y.S.

    2013-01-01

    A simple Boron fiber based interference scheme [1] and other similar schemes are currently routinely used for X-ray coherence estimation at 3rd generation synchrotron radiation sources. If such a scheme is applied after a perfect monochromator and without any focusing / transport optics...... in the optical path, the interpretation of the measured interference pattern is relatively straightforward and can be done in terms of the basic parameters of the source [2]. However, if the interference scheme is used after some focusing optics, e.g. close to the X-ray beam waist, the visibility of fringes can...... be significantly affected by the new shape of the focused beam phase-space. At the same time, optical element imperfections still have a negative impact on the transverse coherence. In such situations, which are frequently encountered in experiments at beamlines, the quantitative interpretation of a measured...

  16. Structure in defocused beams of x-ray mirrors: causes and possible solutions

    Science.gov (United States)

    Sutter, John P.; Alcock, Simon G.; Rust, Fiona; Wang, Hongchang; Sawhney, Kawal

    2014-09-01

    Grazing incidence mirrors are now a standard optic for focusing X-ray beams. Both bimorph and mechanically bendable mirrors are widely used at Diamond Light Source because they permit a wide choice of focal lengths. They can also be deliberately set out of focus to enlarge the X-ray beam, and indeed many beamline teams now wish to generate uniform beam spots of variable size. However, progress has been slowed by the appearance of fine structure in these defocused beams. Measurements showing the relationship between the medium-frequency polishing error and this structure over a variety of beam sizes will be presented. A theoretical model for the simulations of defocused beams from general mirrors will then be developed. Not only the figure error and its first derivative the slope error, but also the second derivative, the curvature error, must be considered. In conclusion, possible ways to reduce the defocused beam structure by varying the actuators' configuration and settings will be discussed.

  17. Active interlock system for high power insertion devices in the x-ray ring

    International Nuclear Information System (INIS)

    Anon.

    1991-01-01

    In modern synchrotron radiation rings photon beams generated by high power insertion devices (IDs) may be sufficiently intense to cause severe thermal damage to the machine vacuum chamber if mis-steered. Thus when operating a storage ring with the IDs, great care must be exercised to prevent mis-steering of the electron beam orbit. At present, the X-ray ring operates with three IDs, namely two hybrid wigglers (HBW) at X-21 and X-25,a nd a 5-Tesla superconducting wiggler at X-17. All wigglers are located in low-beta straight sections. In the straight sections. In the straight sections, beam orbit may be deflected by as much as ±8 mrad without scraping the chamber wall. For various reasons it was not possible to design the X-ray ring vacuum chamber to be safe under all possible operating conditions, however, the chamber is safe for i < 7 mA, all horizontal beam deflection angles and for vertical angles < ± 2.5 mrad. To protect the machine vacuum chamber from damage due to mis-steered beams, and interlock system has been developed and installed. This system utilizes active beam position detectors which continuously monitor beam motion in each of the ID straight sections and logic circuitry which interrupts the RF and dumps the stored beam in the case of a fault

  18. X-ray trace element analysis with positive ion beams

    International Nuclear Information System (INIS)

    Davis, R.H.

    1973-01-01

    A new trace element analysis having the advantage that many elements may be detected in a single measurement, based on positive charged particle induced X-ray florescence and on the production of X-rays by heavy ions, is described. Because of the large cross-sections for the production of discrete X-ray and the low yield of continuum radiation, positive charged particle X-ray florescence is a competitive, fast, analytic tool. In the experiment a beam of positive charged particles from an accelerator was directed toward a target. X-rays induced by the bombardment were detected by a Si(Li) detector the ouput from which was amplified and sorted in a multichannel analyzer. For rapid data handling and analysis, the multichannel analyzer or ADC unit was connected to an on-line computer. A large variety of targets prepared in collaboration with the oceanographers have been studied and spectra obtained for different particles having the same velocity are presented to show that the yield of discrete X-rays increases at least as rapidly as Z 2 . While protons of several MeV appear to be already competitive further advantage may be gained by heavy ions at lower energies since the continuum is reduced while the peak ''signals'' retain strength due to the Z 2 dependence. (S.B.)

  19. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral dental X-ray apparatus for panoramic radiography is described in detail. It comprises a tubular target carrier supporting at its distal end a target with an inclined forward face. Image definition is improved by positioning in the path of the X-rays a window of X-ray transmitting ceramic material, e.g. 90% oxide of Be, or Al, 7% Si0 2 . The target carrier forms a probe which can be positioned in the patient's mouth. X-rays are directed forwardly and laterally of the target to an X-ray film positioned externally. The probe is provided with a detachable sleeve having V-form arms of X-ray opaque material which serve to depress the tongue out of the radiation path and also shield the roof of the mouth and other regions of the head from the X-ray pattern. A cylindrical lead shield defines the X-ray beam angle. (author)

  20. X-ray diagnostic installation for X-ray tomographic images

    International Nuclear Information System (INIS)

    Haendle, J.; Sklebitz, H.

    1984-01-01

    An exemplary embodiment includes at least one x-ray tube for the generation of an x-ray beam, a patient support, an image detector, and a control generator-connected with the x-ray tube and the image detector-for the purpose of moving the x-ray beam, and in opposition thereto, the image field of the image detector. There is connected to the control generator a layer height computer which calculates the enlargement from the geometric data for the tomogram. The image detector has a circuit-connected with the layer height computer-for the purpose of fading-in a marking for the dimensions in the layer plane

  1. Characterization of ion-beam mixed multilayers via grazing x-ray reflectometry

    International Nuclear Information System (INIS)

    Le Boite, M.G.; Traverse, A.; Nevot, L.; Pardo, B.; Corno, J.

    1988-01-01

    The grazing x-ray reflectrometry technique was used as a way to study modifications in metallic multilayers induced by ion-beam irradiation. Due to the high sensitivity of the technique, short-range atomic displacements of an atom A in a layer B can be detected so that the first stages of ion-beam mixing can be investigated. The rate of mixing is measured and the compound A/sub 1-//sub x/B/sub x/ formed at the layers' interfaces is characterized

  2. An X-ray beam position monitor based on the photoluminescence of helium gas

    Science.gov (United States)

    Revesz, Peter; White, Jeffrey A.

    2005-03-01

    A new method for white beam position monitoring for both bend magnet and wiggler synchrotron X-ray radiation has been developed. This method utilizes visible light luminescence generated as a result of ionization by the intense X-ray flux. In video beam position monitors (VBPMs), the luminescence of helium gas at atmospheric pressure is observed through a view port using a CCD camera next to the beam line. The beam position, profile, integrated intensity and FWHM are calculated from the distribution of luminescence intensity in each captured image by custom software. Misalignment of upstream apertures changes the image profile making VBPMs helpful for initial alignment of upstream beam line components. VBPMs can thus provide more information about the X-ray beam than most beam position monitors (BPMs). A beam position calibration procedure, employing a tilted plane-parallel glass plate placed in front of the camera lens, has also been developed. The accuracy of the VBPM system was measured during a bench-top experiment to be better than 1 μm. The He-luminescence-based VBPM system has been operative on three CHESS beam lines (F hard-bend and wiggler, A-line wiggler and G-line wiggler) for about a year. The beam positions are converted to analog voltages and used as feedback signals for beam stabilization. In our paper we discuss details of VBPM construction and describe further results of its performance.

  3. Beam dynamics simulation in the X-ray Compton source

    Energy Technology Data Exchange (ETDEWEB)

    Gladkikh, P.; Karnaukhov, I.; Telegin, Yu.; Shcherbakov, A. E-mail: shcherbakov@kipt.kharkov.ua; Zelinsky, A

    2002-05-01

    At the National Science Center 'Kharkov Institute of Physics and Technology' the X-ray source based on Compton scattering has been developed. The computer code for simulation of electron beam dynamics with taking into account the Compton scattering effect based on Monte Carlo method is described in this report. The first results of computer simulation of beam dynamics with electron-photon interaction, parameters of electron and photon beams are presented. Calculations were carried out with the lattice of synchrotron light source SRS-800 Ukrainian Synchrotron Center.

  4. Perfect-crystal x-ray optics to treat x-ray coherence

    International Nuclear Information System (INIS)

    Yamazaki, Hiroshi; Ishikawa, Tetsuya

    2007-01-01

    X-ray diffraction of perfect crystals, which serve as x-ray monochromator and collimator, modifies coherence properties of x-ray beams. From the time-dependent Takagi-Taupin equations that x-ray wavefields obey in crystals, the reflected wavefield is formulated as an integral transform of a general incident wavefield with temporal and spatial inhomogeneity. A reformulation of rocking-curve profiles from the field solution of the Takagi-Taupin equations allows experimental evaluation of the mutual coherence function of x-ray beam. The rigorous relationship of the coherence functions between before and after reflection clarifies how the coherence is transferred by a crystal. These results will be beneficial to developers of beamline optics for the next generation synchrotron sources. (author)

  5. Optics for coherent X-ray applications

    Energy Technology Data Exchange (ETDEWEB)

    Yabashi, Makina, E-mail: yabashi@spring8.or.jp [RIKEN SPring-8 Center, Kouto 1-1-1, Sayo, Hyogo 679-5148 (Japan); Tono, Kensuke [Japan Synchrotron Radiation Research Institute (JASRI), Kouto 1-1-1, Sayo, Hyogo 679-5198 (Japan); Mimura, Hidekazu [The University of Tokyo, Hongo 7-3-1, Bunkyo-ku, Tokyo 113-8656 (Japan); Matsuyama, Satoshi; Yamauchi, Kazuto [Osaka University, 2-1 Yamada-oka, Suita, Osaka 565-0871 (Japan); Tanaka, Takashi; Tanaka, Hitoshi; Tamasaku, Kenji [RIKEN SPring-8 Center, Kouto 1-1-1, Sayo, Hyogo 679-5148 (Japan); Ohashi, Haruhiko; Goto, Shunji [Japan Synchrotron Radiation Research Institute (JASRI), Kouto 1-1-1, Sayo, Hyogo 679-5198 (Japan); Ishikawa, Tetsuya [RIKEN SPring-8 Center, Kouto 1-1-1, Sayo, Hyogo 679-5148 (Japan)

    2014-08-27

    Developments of optics for coherent X-ray applications and their role in diffraction-limited storage rings are described. Developments of X-ray optics for full utilization of diffraction-limited storage rings (DLSRs) are presented. The expected performance of DLSRs is introduced using the design parameters of SPring-8 II. To develop optical elements applicable to manipulation of coherent X-rays, advanced technologies on precise processing and metrology were invented. With propagation-based coherent X-rays at the 1 km beamline of SPring-8, a beryllium window fabricated with the physical-vapour-deposition method was found to have ideal speckle-free properties. The elastic emission machining method was utilized for developing reflective mirrors without distortion of the wavefronts. The method was further applied to production of diffraction-limited focusing mirrors generating the smallest spot size in the sub-10 nm regime. To enable production of ultra-intense nanobeams at DLSRs, a low-vibration cooling system for a high-heat-load monochromator and advanced diagnostic systems to characterize X-ray beam properties precisely were developed. Finally, new experimental schemes for combinative nano-analysis and spectroscopy realised with novel X-ray optics are discussed.

  6. Device for the collimation of a high-energy beam, in particular a X-ray beam

    International Nuclear Information System (INIS)

    Peyser, L.F.

    1976-01-01

    The design of apertures made of radiation-absorbing material intended for limiting an aperture for a radiation beam of high energy, in particular an X-ray beam is claimed. The apertures are shaped as trapezoids, are held movably, and are adjustable by means of a control device. (UWI) [de

  7. A new cone-beam X-ray CT system with a reduced size planar detector

    International Nuclear Information System (INIS)

    Li Liang; Chen Zhiqiang; Zhang Li; Xing Yuxiang; Kang Kejun

    2006-01-01

    In a traditional cone-beam CT system, the cost of product and computation is very high. The authors propose a transversely truncated cone-beam X-ray CT system with a reduced size detector positioned off-center, in which X-ray beams only cover half of the object. The reduced detector size cuts the cost and the X-ray dose of the CT system. The existing CT reconstruction algorithms are not directly applicable in this new CT system. Hence, the authors develop a BPF-type direct backprojection algorithm. Different from the traditional rebinding methods, our algorithm directly backprojects the pretreated projection data without rebinding. This makes the algorithm compact and computationally more efficient. Finally, some numerical simulations and practical experiments are done to validate the proposed algorithm. (authors)

  8. Evaluation of the Beam Quality of Intraoral X-ray Equipment using Intraoral Standard Films

    International Nuclear Information System (INIS)

    Lee, Sang Sub; Kwon, Hyok Rak; Sim, Woo Hyoun; Oh, Seung Hyoun; Lee, Ji Youn; Jeon, Kug Jin; Kim, Kee Deog; Park, Chang Seo

    2000-01-01

    This study was to evaluate the beam quality of intraoral X-ray equipment used at Yonsei University Dental Hospital (YUDH) using the half value layer (HVL) and the characteristic curve of intraoral standard X-ray film. The study was done using the intraoral X-ray equipment used at each clinical department at YUDH. Aluminum filter was used to determine the HVL. Intraoral standard film was used to get the characteristic curve of each intraoral X-ray equipment. Most of the HVLs of intraoral X-ray equipment were higher than the least recommended thickness, but the REX 601 model used at the operative dentistry department and the X-707 model used at the pediatric dentistry department had HVLs lower than the recommended thickness. The slopes of the characteristic curves of films taken using the PANPAS 601 model and REX 601 model at operative dentistry department, the X-70S model of prosthodontic dentistry department, and the REX 601 model at the student clinic were relatively low. HVL and the characteristic curve of X-ray film can be used to evaluate the beam quality of intraoral X-ray equipment. In order to get the best X-ray films with the least radiation exposure to patients and best diagnostic information in clinical dentistry, X-ray equipment should be managed in the planned and organized fashion.

  9. Interface unit and software of X-ray television automatic machine

    International Nuclear Information System (INIS)

    Molodykh, V.A.; Yamanaev, M.S.

    1983-01-01

    Description of the interface unit and specialized software of X-ray television automatic machine is presented. An algorithm for automatic defect survey, measuring of defect geometric parameters with a successive estimate of control quality in accordance with technical norms is proposed. Experimental investigation results on the quality of welded joints of steel tubes obtained using the above system are summarized

  10. High-intensity laser synchrotron x-ray source

    International Nuclear Information System (INIS)

    Pogorelsky, I.V.

    1995-10-01

    A laser interacting with a relativistic electron beam behaves like a virtual wiggler of an extremely short period equal to half of the laser wavelength. This approach opens a route to relatively compact, high-brightness x-ray sources alternative or complementary to conventional synchrotron light sources. Although not new, the Laser Synchrotron Light Source (LSLS) concept is still waiting for a convincing demonstration. Available at the BNL's Accelerator Test Facility (ATF), a high-brightness electron beam and the high-power C0 2 laser may be used as prototype LSLS brick stones. In a feasible demonstration experiment, 10-GW, 100-ps C0 2 laser beam will be brought to a head-on collision with a 10-ps, 0.5-nC, 70 MeV electron bunch. Flashes of well-collimated, up to 9.36-keV (∼ Angstrom) x-rays of 10-ps pulse duration, with a flux of ∼10 19 photons/sec will be produced via linear Compton backscattering. The x-ray spectrum is tunable proportionally to a variable e-beam energy. A natural short-term extension of the proposed experiment would be further enhancement of the x-ray flux to a 10 21 -10 22 photons/sec level, after the ongoing ATF CO 2 laser upgrade to 1 TW peak power and electron bunch shortening to 3 ps. The ATF LSLS x-ray beamline, exceeding by orders of magnitude the peak fluxes attained at the National Synchrotron Light Source (NSLS) x-ray storage ring, may become attractive for certain users, e.g., for biological x-ray microscopy. In addition, a terawatt CO 2 laser will enable harmonic multiplication of the x-ray spectrum via nonlinear Compton scattering

  11. X-ray beam transfer between hollow fibers for long-distance transport

    International Nuclear Information System (INIS)

    Tanaka, Yoshihito; Matsushita, Ryuki; Shiraishi, Ryutaro; Hasegawa, Takayuki; Ishikawa, Kiyoshi; Sawada, Kei; Kohmura, Yoshiki; Takahashi, Isao

    2016-01-01

    Fiber optics for controlling the x-ray beam trajectory has been examined at the synchrotron facility of SPring-8. Up to now, we have achieved beam deflection by several tens of milli-radian and axis shift of around 75 mm with a 1.5 m-long flexible hollow glass capillary. The achievable beam deflecting angle, axis shift, and timing delay are, in principle, proportional to the length, the square of length and the cube of length, respectively. Thus, for further applications, requiring larger beam shift and pulse delay, longer fibers are indispensable. In order to achieve long-distance transport using the fiber, we thus examined the connection transferring x-rays between fibers in an experimental hutch. The acceptance angle at the input end and the throughput efficiency of the second fiber is consistent with the consideration of the output beam divergence of the first fiber. The enhancement of the transfer efficiency is also discussed for the cases of a closer joint and the use of a refractive lens as a coupler.

  12. X-ray beam transfer between hollow fibers for long-distance transport

    Energy Technology Data Exchange (ETDEWEB)

    Tanaka, Yoshihito, E-mail: tanaka@sci.u-hyogo.ac.jp; Matsushita, Ryuki; Shiraishi, Ryutaro; Hasegawa, Takayuki; Ishikawa, Kiyoshi [Graduate School of Material Science, University of Hyogo, 3-2-1 Kouto, Kamigori-cho, Sayo-gun, Hyogo 678-1297 (Japan); Sawada, Kei; Kohmura, Yoshiki [RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148 (Japan); Takahashi, Isao [Department of Physics, School of Science and Technology, Kwansei Gakuin University, 2-1 Gakuen, Sanda, Hyogo 669-1337 Japan (Japan)

    2016-07-27

    Fiber optics for controlling the x-ray beam trajectory has been examined at the synchrotron facility of SPring-8. Up to now, we have achieved beam deflection by several tens of milli-radian and axis shift of around 75 mm with a 1.5 m-long flexible hollow glass capillary. The achievable beam deflecting angle, axis shift, and timing delay are, in principle, proportional to the length, the square of length and the cube of length, respectively. Thus, for further applications, requiring larger beam shift and pulse delay, longer fibers are indispensable. In order to achieve long-distance transport using the fiber, we thus examined the connection transferring x-rays between fibers in an experimental hutch. The acceptance angle at the input end and the throughput efficiency of the second fiber is consistent with the consideration of the output beam divergence of the first fiber. The enhancement of the transfer efficiency is also discussed for the cases of a closer joint and the use of a refractive lens as a coupler.

  13. Study on the quality assurance of diagnostic X-ray machines and assessment of the absorbed dose to patients

    Science.gov (United States)

    Hassan, G. M.; Rabie, N.; Mustafa, K. A.; Abdel-Khalik, S. S.

    2012-09-01

    Radiation exposure and image quality in X-ray diagnostic radiology provide a clear understanding of the relationship between the radiation dose delivered to a patient and image quality in optimizing medical diagnostic radiology. Because a certain amount of radiation is unavoidably delivered to patients, this should be as low as reasonably achievable. Several X-ray diagnostic machines were used at different medical diagnostic centers in Egypt for studying the beam quality and the dose delivered to the patient. This article studies the factors affecting the beam quality, such as the kilo-volt peak (kVp), exposure time (mSc), tube current (mAs) and the absorbed dose in (μGy) for different examinations. The maximum absorbed dose measured per mAs was 594±239 and 12.5±3.7 μGy for the abdomen and the chest, respectively, while the absorbed dose at the elbow was 18±6 μGy, which was the minimum dose recorded. The compound and expanded uncertainties accompanying these measurements were 4±0.35% and 8±0.7%, respectively. The measurements were done through quality control tests as acceptance procedures.

  14. An experimental evaluation of monochromatic x-ray beam position monitors at diamond light source

    Energy Technology Data Exchange (ETDEWEB)

    Bloomer, Chris, E-mail: chris.bloomer@diamond.ac.uk; Rehm, Guenther; Dolbnya, Igor P. [Diamond Light Source Ltd, Oxfordshire (United Kingdom)

    2016-07-27

    Maintaining the stability of the X-ray beam relative to the sample point is of paramount importance for beamlines and users wanting to perform cutting-edge experiments. The ability to detect, and subsequently compensate for, variations in X-ray beam position with effective diagnostics has multiple benefits: a reduction in commissioning and start-up time, less ‘down-time’, and an improvement in the quality of acquired data. At Diamond Light Source a methodical evaluation of a selection of monochromatic X-ray Beam Position Monitors (XBPMs), using a range of position detection techniques, and from a range of suppliers, was carried out. The results of these experiments are presented, showing the measured RMS noise on the position measurement of each device for a given flux, energy, beam size, and bandwidth. A discussion of the benefits and drawbacks of each of the various devices and techniques is also included.

  15. An experimental evaluation of monochromatic x-ray beam position monitors at diamond light source

    International Nuclear Information System (INIS)

    Bloomer, Chris; Rehm, Guenther; Dolbnya, Igor P.

    2016-01-01

    Maintaining the stability of the X-ray beam relative to the sample point is of paramount importance for beamlines and users wanting to perform cutting-edge experiments. The ability to detect, and subsequently compensate for, variations in X-ray beam position with effective diagnostics has multiple benefits: a reduction in commissioning and start-up time, less ‘down-time’, and an improvement in the quality of acquired data. At Diamond Light Source a methodical evaluation of a selection of monochromatic X-ray Beam Position Monitors (XBPMs), using a range of position detection techniques, and from a range of suppliers, was carried out. The results of these experiments are presented, showing the measured RMS noise on the position measurement of each device for a given flux, energy, beam size, and bandwidth. A discussion of the benefits and drawbacks of each of the various devices and techniques is also included.

  16. LabVIEW control software for scanning micro-beam X-ray fluorescence spectrometer.

    Science.gov (United States)

    Wrobel, Pawel; Czyzycki, Mateusz; Furman, Leszek; Kolasinski, Krzysztof; Lankosz, Marek; Mrenca, Alina; Samek, Lucyna; Wegrzynek, Dariusz

    2012-05-15

    Confocal micro-beam X-ray fluorescence microscope was constructed. The system was assembled from commercially available components - a low power X-ray tube source, polycapillary X-ray optics and silicon drift detector - controlled by an in-house developed LabVIEW software. A video camera coupled to optical microscope was utilized to display the area excited by X-ray beam. The camera image calibration and scan area definition software were also based entirely on LabVIEW code. Presently, the main area of application of the newly constructed spectrometer is 2-dimensional mapping of element distribution in environmental, biological and geological samples with micrometer spatial resolution. The hardware and the developed software can already handle volumetric 3-D confocal scans. In this work, a front panel graphical user interface as well as communication protocols between hardware components were described. Two applications of the spectrometer, to homogeneity testing of titanium layers and to imaging of various types of grains in air particulate matter collected on membrane filters, were presented. Copyright © 2012 Elsevier B.V. All rights reserved.

  17. First dedicated in-beam X-ray measurement in heavy-ion fusion reactions

    Energy Technology Data Exchange (ETDEWEB)

    Berner, C. [Technische Universitaet Muenchen, Lehrstuhl E12 (Germany); RIKEN, Research Group for Superheavy Elements (Japan); Henning, W. [Argonne National Laboratory, Physics Division (United States); RIKEN, Research Group for Superheavy Elements (Japan); Muecher, D.; Gernhaeuser, R.; Hellgartner, S.; Maier, L. [Technische Universitaet Muenchen, Lehrstuhl E12 (Germany); Morita, K.; Morimoto, K.; Kaji, D.; Wakabayashi, Y.; Baba, H. [RIKEN, Research Group for Superheavy Elements (Japan); Lutter, R. [Ludwig-Maximilians-Universitaet, Muenchen (Germany)

    2016-07-01

    We report on an experiment aiming at in-beam X-ray spectroscopy of heavy and superheavy elements (SHE). The goal is to establish K-X-ray spectroscopy as a sensitive tool to identify SHE produced in fusion reactions. SHE, formed after cold or hot fusion, are usually identified via the alpha-decay products, which have to be connected to well-known elements. However, various theories predict spontaneous fission as the dominant decay mode for the daughter nuclides. Additionally, half-lives of these elements are expected to increase to values impeding the identification of SHE solely by their decay. The in-beam identification of the characteristic X-rays would precisely allow to identify the charge number of the produced SHE. Experiments were performed at the RIKEN Nishina Centre for Accelerator based Science by using the gas-filled magnet separator GARIS for superheavy element detection. A high-purity, low-energy planar germanium LEGe-detector was adapted to the GARIS system at the target place for the first time in order to measure the element-characteristic, prompt X-ray emission.

  18. Comparative adoption of cone beam computed tomography and panoramic radiography machines across Australia.

    Science.gov (United States)

    Zhang, A; Critchley, S; Monsour, P A

    2016-12-01

    The aim of the present study was to assess the current adoption of cone beam computed tomography (CBCT) and panoramic radiography (PR) machines across Australia. Information regarding registered CBCT and PR machines was obtained from radiation regulators across Australia. The number of X-ray machines was correlated with the population size, the number of dentists, and the gross state product (GSP) per capita, to determine the best fitting regression model(s). In 2014, there were 232 CBCT and 1681 PR machines registered in Australia. Based on absolute counts, Queensland had the largest number of CBCT and PR machines whereas the Northern Territory had the smallest number. However, when based on accessibility in terms of the population size and the number of dentists, the Australian Capital Territory had the most CBCT machines and Western Australia had the most PR machines. The number of X-ray machines correlated strongly with both the population size and the number of dentists, but not with the GSP per capita. In 2014, the ratio of PR to CBCT machines was approximately 7:1. Projected increases in either the population size or the number of dentists could positively impact on the adoption of PR and CBCT machines in Australia. © 2016 Australian Dental Association.

  19. A split-beam probe-pump-probe scheme for femtosecond time resolved protein X-ray crystallography

    Directory of Open Access Journals (Sweden)

    Jasper J. van Thor

    2015-01-01

    Full Text Available In order to exploit the femtosecond pulse duration of X-ray Free-Electron Lasers (XFEL operating in the hard X-ray regime for ultrafast time-resolved protein crystallography experiments, critical parameters that determine the crystallographic signal-to-noise (I/σI must be addressed. For single-crystal studies under low absorbed dose conditions, it has been shown that the intrinsic pulse intensity stability as well as mode structure and jitter of this structure, significantly affect the crystallographic signal-to-noise. Here, geometrical parameters are theoretically explored for a three-beam scheme: X-ray probe, optical pump, X-ray probe (or “probe-pump-probe” which will allow experimental determination of the photo-induced structure factor amplitude differences, ΔF, in a ratiometric manner, thereby internally referencing the intensity noise of the XFEL source. In addition to a non-collinear split-beam geometry which separates un-pumped and pumped diffraction patterns on an area detector, applying an additional convergence angle to both beams by focusing leads to integration over mosaic blocks in the case of well-ordered stationary protein crystals. Ray-tracing X-ray diffraction simulations are performed for an example using photoactive yellow protein crystals in order to explore the geometrical design parameters which would be needed. The specifications for an X-ray split and delay instrument that implements both an offset angle and focused beams are discussed, for implementation of a probe-pump-probe scheme at the European XFEL. We discuss possible extension of single crystal studies to serial femtosecond crystallography, particularly in view of the expected X-ray damage and ablation due to the first probe pulse.

  20. X-ray beam splitting design for concurrent imaging at hard X-ray FELs and synchrotron facilities

    Energy Technology Data Exchange (ETDEWEB)

    Oberta, P., E-mail: peter.oberta@rigaku.com [Institute of Physics, Academy of Sciences of the Czech Republic, v.v.i., Na Slovance 2, Praha 8, CZ-18221 (Czech Republic); Rigaku, Novodvorská 994, Praha 4, CZ-14221 (Czech Republic); Mokso, R. [Swiss Light Source, Paul Scherrer Institut, Villigen, CH-5232 Villigen (Switzerland)

    2013-11-21

    A new configuration of diffractive–refractive optics for beam splitting is investigated. The set-up can be applied to perform imaging with two beams simultaneously. It brings advantages toward dynamic studies using image guided diffraction or fluorescence spectroscopy. The optimal energy range of operation for the beam-splitter is between 7 keV and 24 keV, reaching best efficiency at an energy of 10 keV. Due to the long focusing distances (several tens of meters) created by the diffractive–refractive optics and the higher refraction efficiency in the softer energy range, the presented set-ups are ideal for hard X-ray FEL sources.

  1. Overview of high intensity x-ray and gamma-ray sources

    International Nuclear Information System (INIS)

    Prestwich, K.R.; Lee, J.R.; Ramirez, J.J.; Sanford, T.W.L.; Agee, F.J.; Frazier, G.B.; Miller, A.R.

    1987-01-01

    The requirements for intense x-ray and gamma-ray sources to simulate the radiation effects from nuclear weapons has led to the development of several types of terawatt-pulsed power systems. One example of a major gamma-ray source is Aurora, a 10-MV, 1.6-MA, 120-ns four-module, electron-beam generator. Recent requirements to improve the dose rate has led to the Aurora upgrade program and to the development of the 20-MV, 800-kA, 40-ns Hermes-III electron-beam accelerator. The Aurora program includes improvements to the pulsed power system and research on techniques to improve the pulse shape of the electron beam. Hermes III will feature twenty 1-MV, 800-kA induction accelerator cavities supplying energy to a magnetically insulated transmission line adder. Hermes III will become operational in 1988. Intense x-ray sources consist of pulsed power systems that operate with 1-MV to 2-MV output voltages and up to 25-TW output powers. These high powers are achieved with either low impedance electron-beam generators or multimodular pulsed power systems. The low-impedance generators have high voltage Marx generators that store the energy and then sequentially transfer this energy to pulse-forming transmission lines with lower and lower impedance until the high currents are reached. In the multimode machines, each module produces 0.7-TW to 4-TW output pulses, and all of the modules are connected together to supply energy to a single diode

  2. X-ray evaluation of residual stress distributions within surface machined layer generated by surface machining and sequential welding

    International Nuclear Information System (INIS)

    Taniguchi, Yuu; Okano, Shigetaka; Mochizuki, Masahito

    2017-01-01

    The excessive tensile residual stress generated by welding after surface machining may be an important factor to cause stress corrosion cracking (SCC) in nuclear power plants. Therefore we need to understand and control the residual stress distribution appropriately. In this study, residual stress distributions within surface machined layer generated by surface machining and sequential welding were evaluated by X-ray diffraction method. Depth directional distributions were also investigated by electrolytic polishing. In addition, to consider the effect of work hardened layer on the residual stress distributions, we also measured full width at half maximum (FWHM) obtained from X-ray diffraction. Testing material was a low-carbon austenitic stainless steel type SUS316L. Test specimens were prepared by surface machining with different cutting conditions. Then, bead-on-plate welding under the same welding condition was carried out on the test specimens with different surface machined layer. As a result, the tensile residual stress generated by surface machining increased with increasing cutting speed and showed nearly uniform distributions on the surface. Furthermore, the tensile residual stress drastically decreased with increasing measurement depth within surface machined layer. Then, the residual stress approached 0 MPa after the compressive value showed. FWHM also decreased drastically with increasing measurement depth and almost constant value from a certain depth, which was almost equal regardless of the machining condition, within surface machined layer in all specimens. After welding, the transverse distribution of the longitudinal residual stress varied in the area apart from the weld center according to machining conditions and had a maximum value in heat affected zone. The magnitude of the maximum residual stress was almost equal regardless of the machining condition and decreased with increasing measurement depth within surface machined layer. Finally, the

  3. An injector for the proposed Berkeley Ultrafast X-Ray Light Source

    International Nuclear Information System (INIS)

    Lidia, Steven; Corlett, John; Pusina, Jan; Staples, John; Zholents, Alexander

    2003-01-01

    Berkeley Lab has proposed to build a recirculating linac based X-ray source for ultra-fast dynamic studies [1]. This machine requires a flat electron beam with a small vertical emittance and large x/y emittance ratio to allow for compression of spontaneous undulator emission of soft and hard x-ray pulses, and a low-emittance, round electron beam for coherent emission of soft x-rays via the FEL process based on cascaded harmonic generation [2]. We propose an injector system consisting of two high gradient high repetition rate photo cathode guns [3] (one for each application), an ∼120 MeV super conducting linear accelerator, a 3rd harmonic cavity for linearization of the longitudinal phase space, and a bunch compressor. We present details of the design and the results of particle tracking studies using several computer codes

  4. Short-wavelength soft-x-ray laser pumped in double-pulse single-beam non-normal incidence

    International Nuclear Information System (INIS)

    Zimmer, D.; Ros, D.; Guilbaud, O.; Habib, J.; Kazamias, S.; Zielbauer, B.; Bagnoud, V.; Ecker, B.; Aurand, B.; Kuehl, T.; Hochhaus, D. C.; Neumayer, P.

    2010-01-01

    We demonstrated a 7.36 nm Ni-like samarium soft-x-ray laser, pumped by 36 J of a neodymium:glass chirped-pulse amplification laser. Double-pulse single-beam non-normal-incidence pumping was applied for efficient soft-x-ray laser generation. In this case, the applied technique included a single-optic focusing geometry for large beam diameters, a single-pass grating compressor, traveling-wave tuning capability, and an optimized high-energy laser double pulse. This scheme has the potential for even shorter-wavelength soft-x-ray laser pumping.

  5. A linear motion machine for soft x-ray interferometry

    International Nuclear Information System (INIS)

    Duarte, R.; Howells, M.R.; Hussain, Z.; Lauritzen, T.; McGill, R.

    1997-07-01

    A Fourier Transform X-ray Spectrometer has been designed and built for use at the Advanced light source at Lawrence Berkeley National Laboratory. The design requires a total rectilinear motion of 15 mm with a maximum pitch error of the stage below ±0.4 μradians, to achieve this the authors chose to build the entire machine as a single monolithic flexure. A hydraulic driver with sliding O-ring seals was developed with the intention to provide motion with a stick-slip position error of less than 0.8 nm at a uniform velocity of 20 μm/sec. The machine is comprised of two pairs of nested linear motion flexures, all explained by means of a theory published earlier by Hathaway. Certain manufacturing errors were successfully corrected by an extra weak-link feature in the monolith frame. The engineering details of all the subsystems of the linear motion machine are described and measured performance reported

  6. X-ray detector array

    International Nuclear Information System (INIS)

    Houston, J.M.

    1980-01-01

    The object of the invention (an ionization chamber X-ray detector array for use with high speed computerised tomographic imaging apparatus) is to reduce the time required to produce a tomographic image. The detector array described determines the distribution of X-ray intensities in one or more flat, coplanar X-ray beams. It comprises three flat anode sheets parallel to the X-ray beam, a plurality of rod-like cathodes between the anodes, a detector gas between the electrodes and a means for applying a potential between the electrodes. Each of the X-ray sources is collimated to give a narrow, planar section of X-ray photons. Sets of X-ray sources in the array are pulsed simultaneously to obtain X-ray transmission data for tomographic image reconstruction. (U.K.)

  7. Thermoluminescence characterization of isolated minerals to identify oranges exposed to γ-ray, e-beam, and X-ray for quarantine applications

    International Nuclear Information System (INIS)

    Deokjo Jo; Joong-Ho Kwon; Bhaskar Sanyal; Bhabha Atomic Research Centre, Mumbai; Ju-Woon Lee

    2015-01-01

    Identification of irradiated fruits is of paramount importance to address the limitation of irradiation technology because of varying national and international regulations. Thermoluminescence (TL) analysis was carried out to identify oranges irradiated with γ-ray, electron beam and X-ray. Non-irradiated samples exhibited background TL signals, but all the irradiated samples showed defined TL glow curves characterized by a prominent peak at 158-163 deg C. Characterizations of the irradiated standard minerals showed that feldspars were the major contributors to the TL emission and stable TL signals revealed a successful detection of irradiated oranges even after a prolonged storage. (author)

  8. Atomic structure of machined semiconducting chips: An x-ray absorption spectroscopy study

    Energy Technology Data Exchange (ETDEWEB)

    Paesler, M.; Sayers, D.

    1988-12-01

    X-ray absorption spectroscopy (XAS) has been used to examine the atomic structure of chips of germanium that were produced by single point diamond machining. It is demonstrated that although the local (nearest neighbor) atomic structure is experimentally quite similar to that of single crystal specimens information from more distant atoms indicates the presence of considerable stress. An outline of the technique is given and the strength of XAS in studying the machining process is demonstrated.

  9. A new high quality X-ray source for Cultural Heritage

    International Nuclear Information System (INIS)

    Walter, Ph.; Variola, A.; Zomer, F.; Jaquet, M.

    2009-01-01

    Compton based photon sources have generated much interest since the rapid advance in laser and accelerator technologies has allowed envisaging their utilisation for ultra-compact radiation sources. These should provide X-ray short pulses with a relatively high average flux. Moreover, the univocal dependence between the scattered photon energy and its angle gives the possibility of obtaining a quasi-monochromatic beam with a simple diaphragm system. For the most ambitious projects the expected performance takes into account a rate of 10 12 - 10 13 photons/s, with an angular divergence of few mrad, an X-ray energy cut-off of few tens of keV and a bandwidth ΔE/E ∼ 1-10 %. Even if the integrated rate cannot compete with synchrotron radiation sources, the cost and the compactness of these Compton based machines make them attractive for a wide spectrum of applications. We explore here the interest of these systems for Cultural Heritage preservation. (authors)

  10. X-ray study of surface layers of tungsten monocrystals after electroerosion machining

    International Nuclear Information System (INIS)

    Aleshina, S.A.; Baranov, Yu.V.; Smirnov, I.S.; Marchuk, A.I.

    1981-01-01

    The presence of polycrystal surface layer, approximately 10 μm thick in subjacent layers and the presence of highly developed block structure which is the result of high-temperature effect of electroerosion machining are detected. Angles of disorientation between blocks, which constitute tens of angular minutes, are evaluated using the method of X-ray topography. According to broadening of profile of X-ray diffraction lines analysis of fine crystal structure of the surface layers is conducted. It is shown that the broadening of diffraction lines is mainly connected with the presence of coherent scat-- tering regions

  11. X-ray Beam Spectral Reconstruction Using Laplace Transform and Attenuation Curves

    International Nuclear Information System (INIS)

    Maeng, Seongjin; Lee, Sang Hoo; Kwon, Dahye; Seo, Jihye; Seo, Kyung Won

    2015-01-01

    As the use of X-ray tubes is widely spread mainly for medical diagnostic purposes or industrial applications, there is increasing demand for accurate and convenient way getting of X-ray beam spectral information. While measurement methods may provide quite accurate spectral information, these methods still require expensive detectors (example: HPGe, High Purity Germanium detector) and some conversion of measurement information into real spectrum. It is concluded that Laplace transform-based spectral reconstruction technique given in equations (1) and (2) works well for a 50-kV X-ray source. In this paper we obtained the attenuation curve by the use of MCNPX simulations. We were able to rebuild the X-ray spectrum of 50 kV through this research by Monte Carlo simulation (fitting parameters, a: 1.2921, b: 0.2342, ν: 0.6190, R-squared: 0.9930)

  12. X-ray and gamma radiography devices

    International Nuclear Information System (INIS)

    Abdul Nassir Ibrahim; Azali Muhammad; Ab. Razak Hamzah; Abd. Aziz Mohamed; Mohamad Pauzi Ismail

    2008-01-01

    When we are using this technique, we also must familiar with the device and instrument that used such as gamma projector, crawler, x-ray tubes and others. So this chapter discussed detailed on device used for radiography work. For the x-ray and gamma, their characteristics are same but the source to produce is a big different. X-ray produced from the machine meanwhile, gamma produce from the source such as Co-60 and IR-192. Both are electromagnetic waves. So, the reader can have some knowledge on what is x-ray tube, discrete x-ray and characteristic x-ray, how the machine works and how to control a machine, what is source for gamma emitter, how to handle the projector and lastly difference between x-ray and gamma. Of course this cannot be with the theory only, so detailed must be learned practically.

  13. In situ surface/interface x-ray diffractometer for oxide molecular beam epitaxy

    Energy Technology Data Exchange (ETDEWEB)

    Lee, J. H.; Freeland, J. W.; Hong, Hawoong, E-mail: hhong@aps.anl.gov [Advanced Photon Source, Argonne National Laboratory, Argonne, Illinois 60439 (United States); Tung, I. C. [Advanced Photon Source, Argonne National Laboratory, Argonne, Illinois 60439 (United States); Department of Materials Science and Engineering, Northwestern University, Evanston, Illinois 60208 (United States); Chang, S.-H.; Bhattacharya, A.; Fong, D. D. [Materials Science Division, Argonne National Laboratory, Argonne, Illinois 60439 (United States)

    2016-01-15

    In situ studies of oxide molecular beam epitaxy by synchrotron x-ray scattering has been made possible by upgrading an existing UHV/molecular beam epitaxy (MBE) six-circle diffractometer system. For oxide MBE growth, pure ozone delivery to the chamber has been made available, and several new deposition sources have been made available on a new 12 in. CF (ConFlat, a registered trademark of Varian, Inc.) flange. X-ray diffraction has been used as a major probe for film growth and structures for the system. In the original design, electron diffraction was intended for the secondary diagnostics available without the necessity of the x-ray and located at separate positions. Deposition of films was made possible at the two diagnostic positions. And, the aiming of the evaporation sources is fixed to the point between two locations. Ozone can be supplied through two separate nozzles for each location. Also two separate thickness monitors are installed. Additional features of the equipment are also presented together with the data taken during typical oxide film growth to illustrate the depth of information available via in situ x-ray techniques.

  14. In situ surface/interface x-ray diffractometer for oxide molecular beam epitaxy

    International Nuclear Information System (INIS)

    Lee, J. H.; Freeland, J. W.; Hong, Hawoong; Tung, I. C.; Chang, S.-H.; Bhattacharya, A.; Fong, D. D.

    2016-01-01

    In situ studies of oxide molecular beam epitaxy by synchrotron x-ray scattering has been made possible by upgrading an existing UHV/molecular beam epitaxy (MBE) six-circle diffractometer system. For oxide MBE growth, pure ozone delivery to the chamber has been made available, and several new deposition sources have been made available on a new 12 in. CF (ConFlat, a registered trademark of Varian, Inc.) flange. X-ray diffraction has been used as a major probe for film growth and structures for the system. In the original design, electron diffraction was intended for the secondary diagnostics available without the necessity of the x-ray and located at separate positions. Deposition of films was made possible at the two diagnostic positions. And, the aiming of the evaporation sources is fixed to the point between two locations. Ozone can be supplied through two separate nozzles for each location. Also two separate thickness monitors are installed. Additional features of the equipment are also presented together with the data taken during typical oxide film growth to illustrate the depth of information available via in situ x-ray techniques

  15. Transmission X-ray mirror

    International Nuclear Information System (INIS)

    Lairson, B.M.; Bilderback, D.H.

    1982-01-01

    Transmission X-ray mirrors have been made from 400 A to 10 000 A thick soap films and have been shown to have novel properties. Using grazing angles of incidence, low energy X-rays were reflected from the front surface while more energetic X-rays were transmitted through the mirror largely unattenuated. A wide bandpass monochromator was made from a silicon carbide mirror followed by a soap film transmission mirror and operated in the white beam at the cornell High Energy Synchrotron Source (CHESS). Bandpasses of ΔE/E=12% to 18% were achieved at 13 keV with peak efficiencies estimated to be between 55% and 75%, respectively. Several wide angle scattering photographs of stretched polyethylene and a phospholipid were obtained in 10 s using an 18% bandpass. (orig.)

  16. Design of x-ray diagnostic beam line for a synchrotron radiation source and measurement results

    Energy Technology Data Exchange (ETDEWEB)

    Garg, Akash Deep, E-mail: akash-deep@rrcat.gov.in; Karnewar, A.K.; Ojha, A.; Shrivastava, B.B.; Holikatti, A.C.; Puntambekar, T.A.; Navathe, C.P.

    2014-08-01

    Indus-2 is a 2.5 GeV synchrotron radiation source (SRS) operational at the Raja Ramanna Centre for Advanced Technology (RRCAT) in India. We have designed, developed and commissioned x-ray diagnostic beam line (X-DBL) at the Indus-2. It is based on pinhole array imaging (8–18 keV). We have derived new equations for online measurements of source position and emission angle with pinhole array optics. Measured values are compared with the measurements at an independent x-ray beam position monitor (staggered pair blade monitor) installed in the X-DBL. The measured values are close to the theoretical expected values within ±12 µm (or ±1.5 μrad) for sufficiently wide range of the beam movements. So, beside the beam size and the beam emittance, online information for the vertical position and angle is also used in the orbit steering. In this paper, the various design considerations of the X-DBL and online measurement results are presented.

  17. X-ray - skeleton

    Science.gov (United States)

    ... this page: //medlineplus.gov/ency/article/003381.htm X-ray - skeleton To use the sharing features on this ... Degenerative bone conditions Osteomyelitis Risks There is low radiation exposure. X-rays machines are set to provide the smallest ...

  18. Quality control assessment of diagnostic x-ray facilities in Ghana

    International Nuclear Information System (INIS)

    Amoako, J.K.; Charles, D.F.; Oppong-Adu, C.; Schandorf, C.

    2009-01-01

    Twenty-three X-ray machines located at 20 different hospitals in Ghana were assessed for quality assurance and control. The radiographic parameters evaluated were tube voltage and current, tube output consistency variation with kilovoltage (kV) and current-time (mAs) factor, exposure time accuracy and beam quality as measured by half-value layer. The photographic parameters assessed were type of films, level of film fogging, film speed, contrast index and film processing temperature. Twenty two of the machines were ∼ 99 % output consistency with standard kVp and mAs. Twelve of the machines had output linearity deviation of less than the acceptable 5 %, while 22 machines were within the accepted kVp deviation of 5 %. The film processing temperature at most hospitals exceeded the required level, due to the absence of air conditioners in the darkrooms. The darkroom quality control at all the facilities was very high. Fogging of films was minimal as indicated by the Base + Fog value of 0.3, speed index of film was ∼ 1.65 and contrast index was comparable to the acceptable value of 1.5. The level of timer accuracy was greater than 95 % for all the X-ray machines with the exception of one. The general quality control status of all the X-ray machines and darkrooms assessed were acceptable and within the quality assurance standards. (au)

  19. Study of hard braking x-ray radiation on the radiation-beam complex ''TEMP''

    International Nuclear Information System (INIS)

    Batrakov, A.B.; Glushko, E.G.; Egorov, A.M.; Zinchenko, A.A.; Litvinenko, V.V.; Lonin, Yu.F.; Ponomarev, A.G.; Rybka, A.V.; Fedotov, S.I.; Uvarov, V.T.

    2015-01-01

    A calculation over of basic parameters of the hard brake x-rayed radiation for the microsecond accelerating of relativistic electronic beam T EMP . Optimization of converters is conducted for these aims. Maximal doses are experimentally got brake x-rayed radiation on beam-radiation complex T EMP . The diagrams of orientation of the brake x-rayed radiation are taken off depending on energies of bunches and forms of electrodes.

  20. An empirical model of diagnostic x-ray attenuation under narrow-beam geometry

    International Nuclear Information System (INIS)

    Mathieu, Kelsey B.; Kappadath, S. Cheenu; White, R. Allen; Atkinson, E. Neely; Cody, Dianna D.

    2011-01-01

    Purpose: The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling. Methods: An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-1.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semilogarithmic (exponential) and linear interpolation]. Results: The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R 2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation. Conclusions: The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry).

  1. An empirical model of diagnostic x-ray attenuation under narrow-beam geometry.

    Science.gov (United States)

    Mathieu, Kelsey B; Kappadath, S Cheenu; White, R Allen; Atkinson, E Neely; Cody, Dianna D

    2011-08-01

    The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling. An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-1.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation]. The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation. The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry).

  2. Do we need X-rays?

    International Nuclear Information System (INIS)

    Sadat, Theo

    2004-01-01

    Do we need X-rays? The answer depends on the cost for 25 kGy sterilization, and the cost for 10 kGy decontamination, compared with the costs of other technologies. In the past, the cost of X-ray was compared with that of Cobalt 60, because treatment by pallet was virtually obsessional. It was considered easier to treat in a Contract Service Center even though several Contract Service Centers equipped with 10 MeV high-power accelerators are in operation, and are profitable. Currently small and reliable accelerators are available, which can be inserted into a production line, i.e. instead of treating a product in its shipment carton or on a pallet, the product can be treated as a unit on the production line. In this case there is no need for extra transport, or intermediate storage facilities, and the product is 'just in time'. So we should compare the cost of treatment, taking all these parameters into account (transport, storage, time lapse, etc.) to find out if X-ray treatment could be competitive. The few studies carried out in the past compared X-ray with Cobalt 60, and the economic result was always in favour of Cobalt due to the poor conversion output. This is still valid today, even if we consider very high-power electron beam accelerators. The performance and cost of e-beam in-line, gamma ray and X-ray will be compared

  3. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... X-rays are a form of radiation like light or radio waves. X-rays pass through most objects, including the body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small ...

  4. The characteristics of epoxy resin cured by {gamma}-ray and E-beam

    Energy Technology Data Exchange (ETDEWEB)

    Nho, Y.C. E-mail: ycnho@kaeri.re.kr; Kang, Phil Hyun; Park, Jong Seok

    2004-10-01

    Epoxy resins are widely used as high-performance thermosetting resins for many industrial applications. In this study, the effect of an electron beam (E-beam) and {gamma}-ray irradiation on the curing of epoxy resins was investigated. Diglycidyl ether of bisphenol-A(DGEBA), diglycidyl ether of bisphenol-F(DGEBF) as epoxy resins, triarylsulfonium hexafluoroantimonate(TASHFA), and triarylsulfonium hexafluorophosphate(TASHFP) as initiators were used in this study. The chemical and mechanical characteristics of irradiated epoxy resins were compared after curing of E-beam and {gamma}-ray irradiation up to 50 kGy in N{sub 2} and air atmosphere. We ascertained the effect of oxygen on the radiation curing of epoxy resin. The thermal properties of cured epoxy were investigated using DMA and TGA. Mechanical properties such as flexural strength were measured. The chemical structures of cured epoxy were characterized by FT-NIR. The gel fraction and the stress at yield of epoxy resins irradiated by E-beam and {gamma}-ray in N{sub 2} atmosphere were also compared with those of epoxy resins irradiated by E-beam and {gamma}-ray in air.

  5. X-ray tube

    International Nuclear Information System (INIS)

    Webley, R.S.

    1975-01-01

    The object of the invention described is to provide an X-ray tube providing a scanned X-ray output which does not require a scanned electron beam. This is obtained by an X-ray tube including an anode which is rotatable about an axis, and a source of a beam of energy, for example an electron beam, arranged to impinge on a surface of the anode to generate X-radiation substantially at the region of incidence on the anode surface. The anode is rotatable about the axis to move the region of incidence over the surface. The anode is so shaped that the rotation causes the region of incidence to move in a predetermined manner relative to fixed parts of the tube so that the generated X-radiation is scanned in a predetermined manner relative to the tube. (UK)

  6. Advanced simulations of x-ray beam propagation through CRL transfocators using ray-tracing and wavefront propagation methods

    DEFF Research Database (Denmark)

    Baltser, Jana; Bergbäck Knudsen, Erik; Vickery, Anette

    2011-01-01

    Compound refractive lenses (CRL) are widely used to manipulate synchrotron radiation beams. Accurate modelling of X-ray beam propagation through individual lenses and through "transfocators" composed of a large number of CRLs is of high importance, since it allows for comprehensive optimization...

  7. X-ray examination apparatus

    NARCIS (Netherlands)

    2000-01-01

    The invention relates to an X-ray apparatus which includes an adjustable X-ray filter. In order to adjust an intensity profile of the X-ray beam, an X-ray absorbing liquid is transported to filter elements of the X-ray filter. Such transport is susceptible to gravitational forces which lead to an

  8. Fabrication of the multilayer beam splitters with large area for soft X-ray laser interferometer

    International Nuclear Information System (INIS)

    Wang Zhanshan; Zhang Zhong; Wang Fengli; Wu Wenjuan; Wang Hongchang; Qin Shuji; Chen Lingyan

    2004-01-01

    The soft X-ray laser Mach-Zehnder interferometer is an important tool to measure the electron densities of a laser-produced plasma near the critical surface. The design of a multilayer beam splitter at 13.9 nm for soft X-ray laser Mach-Zehnder interferometer is completed based on the standard of maximizing product of reflectivity and transmission of the beam splitter. The beam splitters which is Mo/Si multilayers on 10 mm x 10 mm area Si 3 N 4 membrane are fabricated using the magnetron sputtering. The figure error of the beam splitter has reached the deep nanometer magnitude by using optical profiler and the product of reflectivity and transmission measured by synchrotron radiation is up to to 4%. (authors)

  9. Energy dependence evaluation of a ZnO detector for diagnostic X-ray beam

    International Nuclear Information System (INIS)

    Valença, C.P.V.; Silveira, M.A.L.; Macedo, M.A.; Santos, M.A.

    2015-01-01

    In the last decades the international organizations of human health and radiation protection have recommended certain care for using X-ray as a diagnosis tool. The current concern is to avoid any type of radiological accident or overdose to the patient. This can be done assessing the parameters of the X-ray equipment and there are various types of detectors available for that: ionizing chamber, semiconductor devices, etc. These detectors must be calibrated so that they can be used for any energy range and such a procedure is correlated with what is called the energy dependence of the detector. In accordance with the stated requirements of IEC 61267, the standard radiation quality beams and irradiation conditions (RQRs) are the tools and techniques for calibrating diagnostic X-Ray instruments and detectors. The purpose of this work is to evaluate the behavior of the energy dependence of a detector fabricated from a zinc oxide (ZnO) nanofilm. A Pantak industrial X-ray equipment was used to generate the RQR radiation quality beams and test three ZnO detector samples. A 6430 sub-femto-ammeter, Keithley, was used to bias the ZnO detector and simultaneously perform the output readings. The results showed that the ZnO device has some increase in its sensitivity to the ionizing radiation as the X-ray effective energy decreases unlike other types of semiconductor electronic devices typically used as an X-ray detector. We can be concluded that, after calibration, the ZnO device can be used as a diagnostic X-ray detector. (author)

  10. Skin dose from radiotherapy X-ray beams: the influence of energy

    International Nuclear Information System (INIS)

    Butson, M.J.; Metcalfe, P.E.; University of Wollongong, Wollongong, NSW; Mathur, J.N.

    1997-01-01

    Skin-sparing properties of megavoltage photon beams are compromised by electron contamination. Higher energy beams do not necessarily produce lower surface and basal cell layer doses due to this electron contamination. For a 5x5 cm field size the surface doses for 6 MVp and 18 M)p X-ray beams are 10% and 7% of their respective maxima. However, at a field size of 40 x 40cm the percentage surface dose is 42% for both 6 MVp and 18 MVp beams. The introduction of beam modifying devices such as block trays can further reduce the skin-sparing advantages of high energy photon beams. Using a 10 mm perspex block tray, the surface doses for 6 MVp and 18 MVp beams with a 5 x 5 cm field size are 10% and 8%, respectively. At 40 x 40cm, surface doses are 61% and 63% for 6 MVp and 18 MVp beams, respectively. This trend is followed at the basal cell layer depth. At a depth of 1 mm, 18 MVp beam doses are always at least 5% smaller than 6 MVp doses for the same depth at all field sizes when normalized to their respective Dmax values. Results have shown that higher energy photon beams produce a negligible reduction of the delivered dose to the basal cell layer (0.1 mm). Only a small increase in skin sparing is seen at the dermal layer (1 mm), which can be negated by the increased exit dose from an opposing field. (authors)

  11. Nano-fabrication of diffractive optics for soft X-ray and atom beam focusing

    International Nuclear Information System (INIS)

    Rehbein, S.

    2002-01-01

    Nano-structuring processes are described for manufacturing diffractive optics for the condenser-monochromator set-up of the transmission X-ray microscope (TXM) and for the scanning transmission X-ray microscope (STXM) at the BESSY II electron storage ring in Berlin. Furthermore, a process for manufacturing free-standing nickel zone plates for helium atom beam focusing experiments is presented. (author)

  12. High-pressure pair distribution function (PDF) measurement using high-energy focused x-ray beam

    Energy Technology Data Exchange (ETDEWEB)

    Hong, Xinguo, E-mail: xhong@bnl.gov; Weidner, Donald J. [Mineral Physics Institute, Stony Brook University, Stony Brook, NY 11794 (United States); Ehm, Lars [Mineral Physics Institute, Stony Brook University, Stony Brook, NY 11794 (United States); National Synchrotron Light Source II, Brookhaven National Laboratory, Upton, NY 11973 (United States); Zhong, Zhong; Ghose, Sanjit [National Synchrotron Light Source II, Brookhaven National Laboratory, Upton, NY 11973 (United States); Duffy, Thomas S. [Department of Geosciences, Princeton University, Princeton, NJ 08544 (United States)

    2016-07-27

    In this paper, we report recent development of the high-pressure pair distribution function (HP-PDF) measurement technique using a focused high-energy X-ray beam coupled with a diamond anvil cell (DAC). The focusing optics consist of a sagittally bent Laue monochromator and Kirkpatrick-Baez (K–B) mirrors. This combination provides a clean high-energy X-ray beam suitable for HP-PDF research. Demonstration of the HP-PDF technique for nanocrystalline platinum under quasi-hydrostatic condition above 30 GPa is presented.

  13. Performance Test of the Next Generation X-Ray Beam Position Monitor System for The APS Upgrade

    Energy Technology Data Exchange (ETDEWEB)

    Yang, B.; Lee, S.; Westferro, F.; Jaski, Y.; Lenkszus, F.; Sereno, N.; Ramanathan, M.

    2017-03-25

    The Advanced Photon Source is developing its next major upgrade (APS-U) based on the multi-bend achromat lattice. Improved beam stability is critical for the upgrade and will require keeping short-time beam angle change below 0.25 µrad and long-term angle drift below 0.6 µrad. A reliable white x-ray beam diagnostic system in the front end will be a key part of the planned beam stabilization system. This system includes an x-ray beam position monitor (XBPM) based on x-ray fluorescence (XRF) from two specially designed GlidCop A-15 absorbers, a second XBPM using XRF photons from the Exit Mask, and two white beam intensity monitors using XRF from the photon shutter and Compton-scattered photons from the front end beryllium window or a retractable diamond film in windowless front ends. We present orbit stability data for the first XBPM used in the feedback control during user operations, as well as test data from the second XBPM and the intensity monitors. They demonstrate that the XBPM system meets APS-U beam stability requirements.

  14. Uniformity of the soft-x-ray emissions from gold foils irradiated by OMEGA laser beams determined by a two-mirror normal-incidence microscope with multilayer coatings

    International Nuclear Information System (INIS)

    Seely, John F.; Boehly, Thomas; Pien, Gregory; Bradley, David

    1998-01-01

    A two-mirror normal-incidence microscope with multilayer coatings was used to image the soft-x-ray emissions from planar foils irradiated by OMEGA laser beams. The bandpass of the multilayer coatings was centered at a wavelength of 48.3 Angstrom (257-eV energy) and was 0.5 Angstrom wide. Five overlapping OMEGA beams, without beam smoothing, were typically incident on the gold foils. The total energy was 1500 J, and the focused intensity was 6x10 13 Wcm -2 . The 5.8x magnified images were recorded by a gated framing camera at various times during the 3-ns laser pulse. A pinhole camera imaged the x-ray emission in the energy range of >2 keV. On a spatial scale of 10 μm, it was found that the soft-x-ray images at 257 eV were quite uniform and featureless. In contrast, the hard-x-ray images in the energy range of >2 kev were highly nonuniform with numerous features of size 150 μm. copyright 1998 Optical Society of America

  15. Quality control assessment of diagnostic X-ray facilities in Ghana ...

    African Journals Online (AJOL)

    Twenty-three X-ray machines located at 20 different hospitals in Ghana were assessed for quality assurance and control. The radiographic parameters evaluated were tube voltage and current, tube output consistency variation with kilovoltage (kV) and current-time (mAs) factor, exposure time accuracy and beam quality as ...

  16. Time-dependent, x-ray spectral unfolds and brightness temperatures for intense Li+ ion beam-driven hohlraums

    International Nuclear Information System (INIS)

    Fehl, D.L.; Chandler, G.A.; Biggs, F.; Dukart, R.J.; Moats, A.R.; Leeper, R.J.

    1996-01-01

    X-ray-producing hohlraums are being studied as indirect drives for Inertial Confinement Fusion targets. In a 1994 target series on the PBFAII accelerator, cylindrical hohlraum targets were heated by an intense Li + ion beam and viewed by an array of 13 time-resolved, filtered x-ray detectors (XRDs). The UFO unfold code and its suite of auxiliary functions were used extensively in obtaining time- resolved x-ray spectra and radiation temperatures from this diagnostic. UFO was also used to obtain fitted response functions from calibration data, to simulate data from blackbody x-ray spectra of interest, to determine the suitability of various unfolding parameters (e.g., energy domain, energy partition, smoothing conditions, and basis functions), to interpolate the XRD signal traces, and to unfold experimental data. The simulation capabilities of the code were useful in understanding an anomalous feature in the unfolded spectra at low photon energies (≤ 100 eV). Uncertainties in the differential and energy-integrated unfolded spectra were estimated from uncertainties in the data. The time-history of the radiation temperature agreed well with independent calculations of the wall temperature in the hohlraum

  17. Time-dependent, x-ray spectral unfolds and brightness temperatures for intense Li+ ion beam-driven hohlraums

    International Nuclear Information System (INIS)

    Fehl, D.L.; Chandler, G.A.; Biggs, F.; Dukart, R.J.; Moats, A.R.; Leeper, R.J.

    1997-01-01

    X-ray-producing hohlraums are being studied as indirect drives for inertial confinement fusion targets. In a 1994 target series on the PBFAII accelerator, cylindrical hohlraum targets were heated by an intense Li + ion beam and viewed by an array of 13 time-resolved, filtered x-ray detectors (XRDs). The unfold operator (UFO) code and its suite of auxiliary functions were used extensively in obtaining time-resolved x-ray spectra and radiation temperatures from this diagnostic. The UFO was also used to obtain fitted response functions from calibration data, to simulate data from blackbody x-ray spectra of interest, to determine the suitability of various unfolding parameters (e.g., energy domain, energy partition, smoothing conditions, and basis functions), to interpolate the XRD signal traces, and to unfold experimental data. The simulation capabilities of the code were useful in understanding an anomalous feature in the unfolded spectra at low photon energies (≤100 eV). Uncertainties in the differential and energy-integrated unfolded spectra were estimated from uncertainties in the data. The time endash history of the radiation temperature agreed well with independent calculations of the wall temperature in the hohlraum. copyright 1997 American Institute of Physics

  18. A comparison of chemical and ionization dosimetry for high-energy x-ray and electron beams

    International Nuclear Information System (INIS)

    Durocher, J.J.; Boese, H.; Cormack, D.V.; Holloway, A.F.

    1981-01-01

    A comparison was made of ferrous sulfate (Fricke) and ionometric methods for determining the absorbed dose in a phantom irradiated with 4-MV x-rays, 25-MV x-rays, or electron beams having various incident energies between 10 and 32 MeV. Both chemical and ionization instruments were calibrated in a 60 Co beam at a point in water where the absorbed dose had been previously determined. The chemical yield measurements were corrected for spatial variations in dose within the volume of the solution and used to obtain a value of the absorbed dose for each of the x-ray and electron beams. The ratios of G-values required for these determinations were taken from ICRU reports 14 and 21. Ionization instrument readings from three types of commercial ionization chambers were used to obtain alternate values of the absorbed dose for each radiation. C lambda and CE values used in determining these ionization values of dose were also taken from the above ICRU reports. For 4-MV x-rays the values of absorbed dose obtained from chemical measurements agreed to within 0.5% with values obtained from ionization measurements; for 25-MV x-rays the chemical values were about 1% higher than the ionization values; for the electron beams the chemical values were 1%-4% below the ionization values. These discrepancies suggest an inconsistency among the recommended G, C lambda, and CE values similar to that which has been noted by other workers

  19. Synchrotron x-ray microbeam characteristics for x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Iida, Atsuo; Noma, Takashi

    1995-01-01

    X-ray fluorescence analysis using a synchrotron x-ray microprobe has become an indispensable technique for non-destructive micro-analysis. One of the most important parameters that characterize the x-ray microbeam system for x-ray fluorescence analysis is the beam size. For practical analysis, however, the photon flux, the energy resolution and the available energy range are also crucial. Three types of x-ray microbeam systems, including monochromatic and continuum excitation systems, were compared with reference to the sensitivity, the minimum detection limit and the applicability to various types of x-ray spectroscopic analysis. 16 refs., 5 figs

  20. Combined X-ray fluorescence and absorption computed tomography using a synchrotron beam

    International Nuclear Information System (INIS)

    Hall, C

    2013-01-01

    X-ray computed tomography (CT) and fluorescence X-ray computed tomography (FXCT) using synchrotron sources are both useful tools in biomedical imaging research. Synchrotron CT (SRCT) in its various forms is considered an important technique for biomedical imaging since the phase coherence of SR beams can be exploited to obtain images with high contrast resolution. Using a synchrotron as the source for FXCT ensures a fluorescence signal that is optimally detectable by exploiting the beam monochromaticity and polarisation. The ability to combine these techniques so that SRCT and FXCT images are collected simultaneously, would bring distinct benefits to certain biomedical experiments. Simultaneous image acquisition would alleviate some of the registration difficulties which comes from collecting separate data, and it would provide increased information about the sample: functional X-ray images from the FXCT, with the morphological information from the SRCT. A method is presented for generating simultaneous SRCT and FXCT images. Proof of principle modelling has been used to show that it is possible to recover a fluorescence image of a point-like source from an SRCT apparatus by suitably modulating the illuminating planar X-ray beam. The projection image can be successfully used for reconstruction by removing the static modulation from the sinogram in the normal flat and dark field processing. Detection of the modulated fluorescence signal using an energy resolving detector allows the position of a fluorescent marker to be obtained using inverse reconstruction techniques. A discussion is made of particular reconstruction methods which might be applied by utilising both the CT and FXCT data.

  1. Sterilization of Bacillus spores by converted X-rays

    International Nuclear Information System (INIS)

    Ito, Hitoshi; Ohki, Yumi; Watanabe, Yuhei; Sunaga, Hiromi; Ishigaki, Isao

    1991-01-01

    Relative sensitivities of endospores of Bacillus pumilus E601, B. subtilis IAM1069, B. megaterium S31 and B. brevis S5 to gamma-rays, converted X-rays (Bremsstrahlung), and electron beams were examined in order to estimate the conditions in which converted X-rays kill Bacillus spores. The radiation sensitivities to gamma-rays, X-rays and electron beams of each strain dried on glass fiber filter without additives were found to be almost equivalent, and D 10 values were obtained as follows: 1.5-1.6 kGy for B. pumilus, 1.4-1.5 kGy for B. subtilis, 1.9-2.0 kGy for B. megaterium, and 1.6-2.0 kGy for B. brevis. The radiation sensitivities of endospores of each strain to electron beams were slightly lower than those to gamma-rays in the dry condition with additives of 2% peptone + 1% glycerin on glass fiber filter. The increase of radiation resistance in the presence of additives was also observed with X-rays, and it was on an intermediate level between those with gamma-rays and electron beams. In the dry condition using cellulose filter paper, only the radiation resistances of B. megaterium and B. brevis in the presence of additives were increased. (author)

  2. Sagittal x-ray beam deviation at asymmetric inclined diffractors

    Czech Academy of Sciences Publication Activity Database

    Korytár, D.; Hrdý, Jaromír; Artemiev, Nikolai; Ferrari, C.; Freund, A.

    2001-01-01

    Roč. 8, - (2001), s. 1136-1139 ISSN 0909-0495 R&D Projects: GA MŠk OK 305; GA MPO PZ-CH/22 Institutional research plan: CEZ:AV0Z1010914 Keywords : x-ray optics * Si(111) W/grooved crystals * inclined diffraction * out-of-diffraction-plane beams * sagittal focusing Subject RIV: BH - Optics, Masers, Lasers Impact factor: 1.519, year: 2001

  3. Scintillator materials for x-ray detectors and beam monitors

    Czech Academy of Sciences Publication Activity Database

    Martin, T.; Koch, A.; Nikl, Martin

    2017-01-01

    Roč. 42, č. 6 (2017), s. 451-456 ISSN 0883-7694 R&D Projects: GA ČR GA16-15569S Institutional support: RVO:68378271 Keywords : scintillator * X-ray detector * beam monitor * synchrotron * thin film Subject RIV: BM - Solid Matter Physics ; Magnetism OBOR OECD: Condensed matter physics (including formerly solid state physics, supercond.) Impact factor: 5.199, year: 2016

  4. Fluctuations on the X-ray intensity beam using a portable X-ray probe based on {sup 6}LiI(Eu) crystal

    Energy Technology Data Exchange (ETDEWEB)

    Araujo, Geraldo P.; Oliveira, Arno H. [Universidade Federal de Minas Gerais (PCTN/UFMG), Belo Horizonte, MG (Brazil). Programa de Pos-Graduacao em Ciencias e Tecnicas Nucleares; Carneiro, Andre C.; Carneiro, Clemente J.G.; Milian, Felix M.; Velasco, Fermin G., E-mail: fermin@uesc.b [Universidade Estadual de Santa Cruz (CPqCTR/UESC), Ilheus, BA (Brazil). Centro de Pesquisa em Ciencias e Tecnologias das Radiacoes

    2011-07-01

    X-rays are produced by accelerating electrons with a high voltage and allowing them to collide with a metal target. This high voltage presents fluctuations that define peak, minimum, and average voltages. Different voltages are applied to the X-ray tube depending on the radiographic applications. A rectifier circuit converts the alternating high voltage to unidirectional high voltage to accelerate electrons in this tube. The fluctuations on the energy in the electron beam depend on the mode of rectification. Both energy of the electrons and X rays intensity fluctuates. A portable probe built with a {sup 6}LiI(Eu) detector coupled to a 10 m light guide and a Hamamatsu photon counting head H9319 was used to measuring X ray intensities. This system is designed to collect up to 10000 counts in intervals of 10 ms to 1 s. Counts were accumulated in time intervals of 10 ms during 10 s. The system starts the count before activating the X-ray apparatus, which is on during a time interval of 100ms. During this period, counts may overflow in consequence high voltage was adjusted to be 40kV, in order to avoid such a problem. For each of these points dose was measured using an ionization chamber. The objectives of this work are to study fluctuations on the X-ray beam and to calibrate the portable probe for measuring radiation doses. Counting rates measured for each 10 ms presented strong variations due to high voltages fluctuations. Both dose and counting rate when correlated with distances between source and detector followed the inverse square law and presented values of R2 near of unit. A calibration curve of the portable system for dose measurements showed also R2 value near of unity. (author)

  5. Calibration of screen-type X-ray films for electron beams

    International Nuclear Information System (INIS)

    Kobayashi, T.; Sato, Y.; Yoshida, K.; Tateyama, N.; Komori, Y.; Nakabayashi, T.; Oyamada, M.; Nishimura, J.

    2002-01-01

    In order to extract the maximum performance of the screen-type X-ray film for the detection of cascade showers in emulsion chambers, we examined the effect of the thickness of the phosphor screen by irradiation with 200 MeV electron beams. The observed data is in agreement with our analytical prediction of the radiative transfer of photons in a phosphor screen. If we use a combination of the specially prepared screen, HR16B, with a phosphor layer of 400 μm and a green-sensitive X-ray film, HA30, the detection threshold energy of cascade showers can be considerably lowered down to 140 GeV. (author)

  6. Calibration of screen-type X-ray films for electron beams

    CERN Document Server

    Kobayashi, T; Yoshida, K; Tateyama, N; Komori, Y; Nakabayashi, T; Oyamada, M; Nishimura, J

    2002-01-01

    In order to extract the maximum performance of the screen-type X-ray film for the detection of cascade showers in emulsion chambers, we examined the effect of the thickness of the phosphor screen by irradiation with 200 MeV electron beams. The observed data is in agreement with our analytical prediction of the radiative transfer of photons in a phosphor screen. If we use a combination of the specially prepared screen, HR16B, with a phosphor layer of 400 mu m and a green-sensitive X-ray film, HA30, the detection threshold energy of cascade showers can be considerably lowered down to 140 GeV. (author)

  7. Simulations of x-ray speckle-based dark-field and phase-contrast imaging with a polychromatic beam

    Energy Technology Data Exchange (ETDEWEB)

    Zdora, Marie-Christine, E-mail: marie-christine.zdora@diamond.ac.uk [Lehrstuhl für Biomedizinische Physik, Physik-Department & Institut für Medizintechnik, Technische Universität München, 85748 Garching (Germany); Diamond Light Source, Harwell Science and Innovation Campus, Didcot, Oxfordshire OX11 0DE (United Kingdom); Department of Physics & Astronomy, University College London, London WC1E 6BT (United Kingdom); Thibault, Pierre [Department of Physics & Astronomy, University College London, London WC1E 6BT (United Kingdom); Pfeiffer, Franz [Lehrstuhl für Biomedizinische Physik, Physik-Department & Institut für Medizintechnik, Technische Universität München, 85748 Garching (Germany); Zanette, Irene [Lehrstuhl für Biomedizinische Physik, Physik-Department & Institut für Medizintechnik, Technische Universität München, 85748 Garching (Germany); Diamond Light Source, Harwell Science and Innovation Campus, Didcot, Oxfordshire OX11 0DE (United Kingdom)

    2015-09-21

    Following the first experimental demonstration of x-ray speckle-based multimodal imaging using a polychromatic beam [I. Zanette et al., Phys. Rev. Lett. 112(25), 253903 (2014)], we present a simulation study on the effects of a polychromatic x-ray spectrum on the performance of this technique. We observe that the contrast of the near-field speckles is only mildly influenced by the bandwidth of the energy spectrum. Moreover, using a homogeneous object with simple geometry, we characterize the beam hardening artifacts in the reconstructed transmission and refraction angle images, and we describe how the beam hardening also affects the dark-field signal provided by speckle tracking. This study is particularly important for further implementations and developments of coherent speckle-based techniques at laboratory x-ray sources.

  8. Carbon-ion beam irradiation kills X-ray-resistant p53-null cancer cells by inducing mitotic catastrophe.

    Directory of Open Access Journals (Sweden)

    Napapat Amornwichet

    Full Text Available BACKGROUND AND PURPOSE: To understand the mechanisms involved in the strong killing effect of carbon-ion beam irradiation on cancer cells with TP53 tumor suppressor gene deficiencies. MATERIALS AND METHODS: DNA damage responses after carbon-ion beam or X-ray irradiation in isogenic HCT116 colorectal cancer cell lines with and without TP53 (p53+/+ and p53-/-, respectively were analyzed as follows: cell survival by clonogenic assay, cell death modes by morphologic observation of DAPI-stained nuclei, DNA double-strand breaks (DSBs by immunostaining of phosphorylated H2AX (γH2AX, and cell cycle by flow cytometry and immunostaining of Ser10-phosphorylated histone H3. RESULTS: The p53-/- cells were more resistant than the p53+/+ cells to X-ray irradiation, while the sensitivities of the p53+/+ and p53-/- cells to carbon-ion beam irradiation were comparable. X-ray and carbon-ion beam irradiations predominantly induced apoptosis of the p53+/+ cells but not the p53-/- cells. In the p53-/- cells, carbon-ion beam irradiation, but not X-ray irradiation, markedly induced mitotic catastrophe that was associated with premature mitotic entry with harboring long-retained DSBs at 24 h post-irradiation. CONCLUSIONS: Efficient induction of mitotic catastrophe in apoptosis-resistant p53-deficient cells implies a strong cancer cell-killing effect of carbon-ion beam irradiation that is independent of the p53 status, suggesting its biological advantage over X-ray treatment.

  9. CRL X-ray tube

    International Nuclear Information System (INIS)

    Kolchevsky, N.N.; Petrov, P.V.

    2015-01-01

    A novel types of X-ray tubes with refractive lenses are proposed. CRL-R X-ray tube consists of Compound Refractive Lens- CRL and Reflection X-ray tube. CRL acts as X-ray window. CRL-T X-ray consists of CRL and Transmission X-ray tube. CRL acts as target for electron beam. CRL refractive lens acts as filter, collimator, waveguide and focusing lens. Properties and construction of the CRL X-ray tube are discussed. (authors)

  10. High-current Rhodotron for X-ray facility

    International Nuclear Information System (INIS)

    Umezu, Toru; Tsujiura, Yuichiro; Bol, Jean Louis

    2009-01-01

    The Rhodotron is a widely employed high-power industrial accelerator developed and exclusively distributed by IBA. Most early examples of the accelerator were optimized to operate at 10 MeV. A new Rhodotron configuration recently advanced produces a lower-energy higher-current beam dedicated with x-ray to sterilize and enhancement materials. Core elements of this system's evolution include a higher performance RF electron gun (operating range, response control, and cathode lifetime). This operational machine is now producing 100 mA at 7 MeV (700 kW of beam) and treat medical devices, thick cable and pipes with a high efficiency. (author)

  11. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral dental X-ray apparatus for panoramic dental radiography is described in detail. It comprises an electron gun having an elongated tubular target carrier extending into the patient's mouth. The carrier supports an inclined target for direction of an X-ray pattern towards a film positioned externally of the patient's mouth. Image definition is improved by a focusing anode which focuses the electron beam into a sharp spot (0.05 to 0.10 mm diameter) on the target. The potential on the focusing anode is adjustable to vary the size of the spot. An X-ray transmitting ceramic (oxides of Be, Al and Si) window is positioned adjacent to the front face of the target. The electron beam can be magnetically deflected to change the X-ray beam direction. (author)

  12. Application of the entropy concept to describe the quality of a X-ray beam

    International Nuclear Information System (INIS)

    Oliveira, A.D.

    2002-01-01

    In a previous work (Oliveira and Pedroso de Lima, 2000), we introduced the concept of entropy to describe the degradation of the energy of a photon beam incident in a material. We pointed out the relation between both, the entropy of the distribution of the energy deposited in matter and the traditional quality factor of the radiation protection. From the point of view of the entropy, we showed that, in what concerns to the monoenergetic approximation to a polyenergetic beam, we can hardly speak in an approximation (Oliveira 2001). In fact, the behaviour of the entropy is very different for both, an X-ray and a monoenergetic beam. In this work we developed further the study of the degradation of the energy of X-ray photons, from the point of view of the entropy, introducing the concepts of surface entropy, S surf , equilibrium entropy, S eq , and concepts of entropy variation, namely, primary to surface variation, S ps , and total entropy variation, S T . These are characteristic parameters of the X-ray beams describing the degradation of the primary photons while they interact with matter

  13. DCARR: a spectrograph for measuring low-energy x rays

    International Nuclear Information System (INIS)

    Anon.

    1978-01-01

    DCARR, the Differential Critical Angle Reflection Refraction detector system, is described. This detector was designed to measure low-energy x rays, 500 to 5000 eV, with a high degree of resolution, 250 eV. DCARR was developed because these low-energy measurements are of interest in the diagnostics of x-radiation in nuclear tests and available equipment could not make measurements at this low an energy in field tests. DCARR is a versatile piece of equipment that can also be used as a laboratory tool, such as in measuring the low-energy x rays emitted by lasers and various x-ray machines

  14. Electron beam requirements for soft x-ray/XUV free-electron lasers

    International Nuclear Information System (INIS)

    Goldstein, J.C.

    1987-01-01

    A discussion of the electron beam quality (peak current, energy spread, and transverse emittance) required to drive short wavelength free-electron lasers in the XUV (10-100 nm) and soft x-ray (<10 nm) optical wavelength ranges is presented

  15. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  16. Modern X-ray difraction. X-ray diffractometry for material scientists, physicists, and chemicists

    International Nuclear Information System (INIS)

    Spiess, L.; Schwarzer, R.; Behnken, H.; Teichert, G.

    2005-01-01

    The book yields a comprehensive survey over the applications of X-ray diffraction in fields like material techniques, metallurgy, electrotechniques, machine engineering, as well as micro- and nanotechniques. The necessary fundamental knowledge on X-ray diffraction are mediated foundedly and illustratively. Thereby new techniques and evaluation procedures are presented as well as well known methods. The content: Production and properties of X radiation, diffraction of X radiation, hardware for X-ray diffraction, methods of X-ray diffraction, lattice-constant determination, phase analysis, X-ray profile analysis, crystal structure analysis, X-ray radiographic stress analysis, X-ray radiographic texture analysis, crystal orientation determination, pecularities at thin films, small angle scattering

  17. Glass capillary optics for making x-ray beams of 0.1 to 50 microns diameter

    International Nuclear Information System (INIS)

    Bilderback, Donald H.; Fontes, Ernest

    1997-01-01

    We have fabricated a unique computerized glass puller that can make parabolic or elliptically tapered glass capillaries for microbeam x-ray experiments from hollow glass tubing. We have produced optics that work in a single-bounce imaging mode or in a multi-bounce condensing mode. The imaging-mode capillaries have been used to create 20 to 50 micron diameter x-ray beams at 12 keV that are quite useful for imaging diffraction patterns from tiny bundles of carbon and Kevlar fibers. The condensing-mode capillaries are useful for creating submicron diameter beams and show great promise in x-ray fluorescence applications with femtogram sensitivity for patterned Er and Ti dopants diffused into an optically-active lithium niobate wafer

  18. Photoneutron intensity variation with field size around radiotherapy linear accelerator 18-MeV X-ray beam

    Energy Technology Data Exchange (ETDEWEB)

    Al-Ghamdi, H.; Fazal-ur-Rehman [Physics Department, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia); Al-Jarallah, M.I. [Physics Department, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia)], E-mail: mibrahim@kfupm.edu.sa; Maalej, N. [Physics Department, King Fahd University of Petroleum and Minerals, Dhahran 31261 (Saudi Arabia)

    2008-08-15

    In X-ray radiotherapy accelerators, neutrons are produced mainly by ({gamma},n) reaction when high energy X-rays interact with high Z materials of the linear accelerator head. These materials include the lead (Pb) used as shielding in the collimator, tungsten (W) target used for the production of X-rays and iron (Fe) in the accelerator head. These unwanted neutrons contaminate the therapeutic beam and contribute to the patient dose during the treatment of a cancer patient. Knowing the neutron distribution around the radiotherapy accelerator is therefore desired. CR-39 nuclear track detectors (NTDs) were used to study the variation of fast and thermal neutron relative intensities around an 18 MeV linear accelerator X-ray beam with the field sizes of 0, 10x10, 20x20, 30x30 and 40x40cm{sup 2}. For fast neutron detection, bare NTDs were used. For thermal neutron detection, NTDs were covered with lithium tetra borate (Li{sub 2}B{sub 4}O{sub 7}) converters. The NTDs were placed at different locations in the direction perpendicular to the treatment couch (transversal) and in the direction parallel to the treatment couch (longitudinal) with respect to the isocenter of the accelerator. The fast neutron relative intensity is symmetrical about the beam axis and exhibits an exponential-like drop with distance from the isocenter of the accelerator for all the field sizes. At the primary beam (isocenter), the relative fast neutron intensity is highest for 40x40cm{sup 2} field size and decreases linearly with the decrease in the field size. However, fast neutron intensities do not change significantly with beam size for the measurements outside the primary beam. The fast neutron intensity in the longitudinal direction outside the primary beam decreases linearly with the field size. The thermal neutron intensity, at any location, was found to be almost independent of the field size.

  19. Discovery of Hard Nonthermal Pulsed X-Ray Emission from the Anomalous X-Ray Pulsar 1E 1841-045

    NARCIS (Netherlands)

    Kuiper, L.; Hermsen, W.; Méndez, R.M.

    2004-01-01

    We report the discovery of nonthermal pulsed X-ray/soft gamma-ray emission up to ~150 keV from the anomalous 11.8 s X-ray pulsar AXP 1E 1841-045 located near the center of supernova remnant Kes 73 using Rossi X-Ray Timing Explorer (RXTE) Proportional Counter Array and High Energy X-Ray Timing

  20. Compact X-ray sources: X-rays from self-reflection

    Science.gov (United States)

    Mangles, Stuart P. D.

    2012-05-01

    Laser-based particle acceleration offers a way to reduce the size of hard-X-ray sources. Scientists have now developed a simple scheme that produces a bright flash of hard X-rays by using a single laser pulse both to generate and to scatter an electron beam.

  1. Mode-Locked Multichromatic X-Rays in a Seeded Free-Electron Laser for Single-Shot X-Ray Spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Xiang, Dao; Ding, Yuantao; Raubenheimer, Tor; Wu, Juhao; /SLAC

    2012-05-10

    We present the promise of generating gigawatt mode-locked multichromatic x rays in a seeded free-electron laser (FEL). We show that, by using a laser to imprint periodic modulation in electron beam phase space, a single-frequency coherent seed can be amplified and further translated to a mode-locked multichromatic output in an FEL. With this configuration the FEL output consists of a train of mode-locked ultrashort pulses which span a wide frequency gap with a series of equally spaced sharp lines. These gigawatt multichromatic x rays may potentially allow one to explore the structure and dynamics of a large number of atomic states simultaneously. The feasibility of generating mode-locked x rays ranging from carbon K edge ({approx}284 eV) to copper L{sub 3} edge ({approx}931 eV) is confirmed with numerical simulation using the realistic parameters of the linac coherent light source (LCLS) and LCLS-II. We anticipate that the mode-locked multichromatic x rays in FELs may open up new opportunities in x-ray spectroscopy (i.e. resonant inelastic x-ray scattering, time-resolved scattering and spectroscopy, etc.).

  2. Providing Bright-Hard X-ray Beams from a Lower Energy Light Source

    Science.gov (United States)

    Robin, David

    2002-04-01

    At the Advanced Light Source (ALS) there had been an increasing demand for more high brightness harder X-ray sources in the 7 to 40 KeV range. In response to that demand, the ALS storage ring was modified in August 2001. Three 1.3 Tesla normal conducting bending magnets were removed and replaced with three 5 Tesla superconducting magnets (Superbends). The radiation produced by these Superbends is an order of magnitude higher in photon brightness and flux at 12 keV than the 1.3 Tesla bends, making them excellent sources of harder x-rays for protein crystallography and other harder x-ray applications. At the same time the Superbends do not compromise the performance of the facility in the UV and Soft X-ray regions of the spectrum. The Superbends will eventually feed 12 new x-ray beam lines greatly enhancing the facility's capacity in the hard x-ray region. The Superbend project is the biggest upgrade to the ALS storage ring since the ring was commissioned in 1993. In this paper we present, a history of the project, details of the magnet, installation, commissioning, and resulting performance of the ALS with Superbends.

  3. Transmission of broad W/Rh and W/Al (target/filter) x-ray beams operated at 25-49 kVp through common shielding materials.

    Science.gov (United States)

    Li, Xinhua; Zhang, Da; Liu, Bob

    2012-07-01

    To provide transmission data for broad 25-39 kVp (kilovolt peak) W/Rh and 25-49 kVp W/Al (target/filter, W-tungsten, Rh-rhodium, and Al-aluminum) x-ray beams through common shielding materials, such as lead, concrete, gypsum wallboard, wood, steel, and plate glass. The unfiltered W-target x-ray spectra measured on a Selenia Dimensions system (Hologic Inc., Bedford, MA) set at 20-49 kVp were, respectively, filtered using 50-μm Rh and 700-μm Al, and were subsequently used for Monte Carlo calculations. The transmission of broad x-ray beams through shielding materials was simulated using Geant4 low energy electromagnetic physics package with photon- and electron-processes above 250 eV, including photoelectric effect, Compton scattering, and Rayleigh scattering. The calculated transmission data were fitted using Archer equation with a robust fitting algorithm. The transmission of broad x-ray beams through the above-mentioned shielding materials was calculated down to about 10(-5) for 25-39 kVp W/Rh and 25-49 kVp W/Al. The fitted results of α, β, and γ in Archer equation were provided. The α values of kVp ≥ 40 were approximately consistent with those of NCRP Report No. 147. These data provide inputs for the shielding designs of x-ray imaging facilities with W-anode x-ray beams, such as from Selenia Dimensions.

  4. Errors in dual x-ray beam differential absorptiometry

    International Nuclear Information System (INIS)

    Bolin, F.; Preuss, L.; Gilbert, K.; Bugenis, C.

    1977-01-01

    Errors pertinent to the dual beam absorptiometry system have been studied and five areas are given in detail: (1) scattering, in which a computer analysis of multiple scattering shows little error due to this effect; (2) geometrical configuration effects, in which the slope of the sample is shown to influence the accuracy of the measurement; (3) Poisson variations, wherein it is shown that a simultaneous reduction can be obtained in both dosage and statistical error; (4) absorption coefficients, in which the effect of variation in absorption coefficient compilations is shown to have a critical effect on the interpretations of experimental data; and (5) filtering, wherein is shown the need for filters on dual beam systems using a characteristic x-ray output. A zero filter system is outlined

  5. Detection of coherent X-ray transition radiation and its application to beam diagnostics

    International Nuclear Information System (INIS)

    Piestrup, M.A.; Boyers, D.G.; Pincus, C.I.; Li Qiang; Moran, M.J.; Bergstrom, J.C.; Caplan, H.S.; Silzer, R.M.; Skopik, D.M.; Rothbart, G.B.

    1989-01-01

    We investigate the use of coherent X-ray transition radiation to measure the energy of ultra-relativistic charged particles. This can be used for beam diagnostics for both high-repetition-rate and single-pulse, high-current accelerators. The research also has possible applications for the detection and identification of these particles. By selecting foil thickness and spacing, it is possible to design radiators whose angle of emission varies radically over a range of charged particle energies. We have constructed three coherent radiators and tested them at two accelerators using electron beam energies ranging from 50 to 228 MeV. Soft X-ray emission (1-3 keV) was emitted in a circularly symmetrical annulus with half-angle divergence of 2.5-9.0 mrad. The angle of peak emission was found to increase with electron-beam energy, in contrast to the incoherent case for which the angle of emission varies inversely with electron-beam energy. (orig.)

  6. The exponential edge-gradient effect in x-ray computed tomography

    International Nuclear Information System (INIS)

    Joseph, P.M.

    1981-01-01

    The exponential edge-gradient effect must arise in any X-ray transmission CT scanner whenever long sharp edges of high contrast are encountered. The effect is non-linear and is due to the interaction of the exponential law of X-ray attenuation and the finite width of the scanning beam in the x-y plane. The error induced in the projection values is proved to be always negative. While the most common effect is lucent streaks emerging from single straight edges, it is demonstrated that dense streaks from pairs of edges are possible. It is shown that an exact correction of the error is possible only under very special (and rather unrealistic) circumstances in which an infinite number of samples per beam width are available and all thin rays making up the beam can be considered parallel. As a practical matter, nevertheless, increased sample density is highly desirable in making good approximate corrections; this is demonstrated with simulated scans. Two classes of approximate correction algorithms are described and their effectiveness evaluated on simulated CT phantom scans. One such algorithm is also shown to work well with a real scan of a physical phantom on a machine that provides approximately four samples per beam width. (author)

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... wrist, arm, elbow, shoulder, spine, pelvis, hip, thigh, knee, leg (shin), ankle or foot. top of page ... the patient standing upright, as in cases of knee x-rays. A portable x-ray machine is ...

  8. 3D algebraic iterative reconstruction for cone-beam x-ray differential phase-contrast computed tomography.

    Science.gov (United States)

    Fu, Jian; Hu, Xinhua; Velroyen, Astrid; Bech, Martin; Jiang, Ming; Pfeiffer, Franz

    2015-01-01

    Due to the potential of compact imaging systems with magnified spatial resolution and contrast, cone-beam x-ray differential phase-contrast computed tomography (DPC-CT) has attracted significant interest. The current proposed FDK reconstruction algorithm with the Hilbert imaginary filter will induce severe cone-beam artifacts when the cone-beam angle becomes large. In this paper, we propose an algebraic iterative reconstruction (AIR) method for cone-beam DPC-CT and report its experiment results. This approach considers the reconstruction process as the optimization of a discrete representation of the object function to satisfy a system of equations that describes the cone-beam DPC-CT imaging modality. Unlike the conventional iterative algorithms for absorption-based CT, it involves the derivative operation to the forward projections of the reconstructed intermediate image to take into account the differential nature of the DPC projections. This method is based on the algebraic reconstruction technique, reconstructs the image ray by ray, and is expected to provide better derivative estimates in iterations. This work comprises a numerical study of the algorithm and its experimental verification using a dataset measured with a three-grating interferometer and a mini-focus x-ray tube source. It is shown that the proposed method can reduce the cone-beam artifacts and performs better than FDK under large cone-beam angles. This algorithm is of interest for future cone-beam DPC-CT applications.

  9. Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype.

    Science.gov (United States)

    Speidel, Michael A; Tomkowiak, Michael T; Raval, Amish N; Dunkerley, David A P; Slagowski, Jordan M; Kahn, Paul; Ku, Jamie; Funk, Tobias

    Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system for low dose cardiac imaging. The use of a narrow scanned x-ray beam in SBDX reduces detected x-ray scatter and improves dose efficiency, however the tight beam collimation also limits the maximum achievable x-ray fluence. To increase the fluence available for imaging, we have constructed a new SBDX prototype with a wider x-ray beam, larger-area detector, and new real-time image reconstructor. Imaging is performed with a scanning source that generates 40,328 narrow overlapping projections from 71 × 71 focal spot positions for every 1/15 s scan period. A high speed 2-mm thick CdTe photon counting detector was constructed with 320×160 elements and 10.6 cm × 5.3 cm area (full readout every 1.28 μs), providing an 86% increase in area over the previous SBDX prototype. A matching multihole collimator was fabricated from layers of tungsten, brass, and lead, and a multi-GPU reconstructor was assembled to reconstruct the stream of captured detector images into full field-of-view images in real time. Thirty-two tomosynthetic planes spaced by 5 mm plus a multiplane composite image are produced for each scan frame. Noise equivalent quanta on the new SBDX prototype measured 63%-71% higher than the previous prototype. X-ray scatter fraction was 3.9-7.8% when imaging 23.3-32.6 cm acrylic phantoms, versus 2.3-4.2% with the previous prototype. Coronary angiographic imaging at 15 frame/s was successfully performed on the new SBDX prototype, with live display of either a multiplane composite or single plane image.

  10. Development of a Method to Assess the Precision Of the z-axis X-ray Beam Collimation in a CT Scanner

    Science.gov (United States)

    Kim, Yon-Min

    2018-05-01

    Generally X-ray equipment specifies the beam collimator for the accuracy measurement as a quality control item, but the computed tomography (CT) scanner with high dose has no collimator accuracy measurement item. If the radiation dose is to be reduced, an important step is to check if the beam precisely collimates at the body part for CT scan. However, few ways are available to assess how precisely the X-ray beam is collimated. In this regard, this paper provides a way to assess the precision of z-axis X-ray beam collimation in a CT scanner. After the image plate cassette had been exposed to the X-ray beam, the exposed width was automatically detected by using a computer program developed by the research team to calculate the difference between the exposed width and the imaged width (at isocenter). The result for the precision of z-axis X-ray beam collimation showed that the exposed width was 3.8 mm and the overexposure was high at 304% when a narrow beam of a 1.25 mm imaged width was used. In this study, the precision of the beam collimation of the CT scanner, which is frequently used for medical services, was measured in a convenient way by using the image plate (IP) cassette.

  11. Development of isodose curves for A 6 mv x-ray beam

    International Nuclear Information System (INIS)

    Hidaytalla, Lamya Abbas

    1994-08-01

    In this thesis radiation distribution of 6 Mv x-ray beam in a water phantom is developed. The method is based on a simple empirical equation and the assumption that the x-ray source is a point source. This leads to a simple equation for the calculation and plotting of the isodose curves. the charts obtained for two fields, 6*6 and 12*12 cm 2 show good agreement with the previous data used in the isotope and radiation centre, Khartoum hospital. It is suggested that further development should be carried out by writing a computer program for all the fields. (Author)

  12. On beam quality and stopping power ratios for high-energy x-rays

    International Nuclear Information System (INIS)

    Johnsson, S.A.; Ceberg, C.P.; Knoeoes, T.; Nilsson, P.

    2000-01-01

    The aim of this work is to quantitatively compare two commonly used beam quality indices, TPR(20/10) and %dd(10) x , with respect to their ability to predict stopping power ratios (water to air), s w,air , for high-energy x-rays. In particular, effects due to a varied amount of filtration of the photon beam will be studied. A new method for characterizing beam quality is also presented, where the information we strive to obtain is the moments of the spectral distribution. We will show how the moments enter into a general description of the transmission curve and that it is possible to correlate the moments to s w,air with a unique and simple relationship. Comparisons with TPR(20/10) and %dd(10) x show that the moments are well suited for beam quality specification in terms of choosing the correct s w,air . (author)

  13. Correlation of electron beams and hard x-ray emissions in ISTTOK Tokamak

    Energy Technology Data Exchange (ETDEWEB)

    Jakubowski, L.; Malinowski, K.; Sadowski, M.J.; Zebrowski, J.; Rabinski, M.; Jakubowski, M.J. [National Centre for Nuclear Research (NCBJ), Otwock (Poland); Plyusnin, V.V.; Fernandes, H.; Silva, C.; Duarte, P. [Association Euratom/IST, Instituto de Plasmas e Fusao Nuclear, Instituto Superior Tecnico, Lisboa (Portugal)

    2013-11-15

    The paper reports on experimental studies of electron beams in the ISTTOK tokamak, those were performed by means of an improved four-channel detector. The Cherenkov-type detector measuring head was equipped with four radiators made of two types of alumina-nitrate (AlN) poly-crystals: machinable and translucent ones, both of 10 mm in diameter and 2.5 mm in thickness. The movable support that enabled the whole detectors to be placed inside the tokamak vacuum chamber, at chosen positions along the ISTTOK minor radius. Since the electron energy distribution is one of the most important characteristics of tokamak plasmas, the main aim of the study was to perform estimations of an energy spectrum of the recorded electrons. For this purpose the radiators were coated with molybdenum (Mo) layers of different thickness. The technique based on the use of Cherenkov-type detectors enabled the detection of fast electrons (of energy above 66 keV) and determination of their spatial and temporal characteristics in the ISTTOK experiment. Measurements of hard X-rays (HXR), which were emitted during ISTTOK discharges, have also been performed. Particular attention was paid to the correlation measurements of HXR pulses with run-away electron beams. (copyright 2013 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  14. Simulations of X-ray synchrotron beams using the EGS4 code system in medical applications

    International Nuclear Information System (INIS)

    Orion, I.; Henn, A.; Sagi, I.; Dilmanian, F.A.; Pena, L.; Rosenfeld, A.B.

    2001-01-01

    X-ray synchrotron beams are commonly used in biological and medical research. The availability of intense, polarized low-energy photons from the synchrotron beams provides a high dose transfer to biological materials. The EGS4 code system, which includes the photoelectron angular distribution, electron motion inside a magnetic field, and the LSCAT package, found to be the appropriate Monte Carlo code for synchrotron-produced X-ray simulations. The LSCAT package was developed in 1995 for the EGS4 code to contain the routines to simulate the linear polarization, the bound Compton, and the incoherent scattering functions. Three medical applications were demonstrated using the EGS4 Monte Carlo code as a proficient simulation code system for the synchrotron low-energy X-ray source. (orig.)

  15. Multilayer beam splitter used in a soft X-ray Mach-Zehnder interferometer at working wavelength of 13.9 nm

    International Nuclear Information System (INIS)

    Zhang Zhong; Wang Zhanshan; Wang Hongchang; Wang Fengli; Wu Wenjuan; Zhang Shumin; Qin Shuji; Chen Lingyan

    2006-01-01

    The soft X-ray Mach-Zehnder interferometer is an important tool in measuring the electron densities of laser-produced plasma near the critical surface. The design, fabrication and characterization of multilayer beam splitters at 13.9 nm for soft X-ray Mach-Zehnder interferometer are presented in the paper. The design of beam splitter is completed based on the standard of maximizing product of reflectivity and transmission of the beam splitter at 13.9 nm. The beam splitters, which are Mo/Si multi-layer deposited on 10 mm x 10 mm area, 100 nm thickness Si 3 N 4 membranes, are fabricated using the magnetron sputtering. A method based on extended He-Ne laser beam is developed to analyze the figure error of the beam splitters. The data measured by an optical profiler prove that the method based on visible light is effective to analyze the figure of the beam splitters. The rms figure error of a beam splitter reaches 1.757 nm in the center area 3.82 mm x 3.46 mm and satisfies the need of soft X-ray interference experiment. The product of reflectivity and transmission measured by synchrotron radiation is near to 4%. The Mach-Zehnder interferometer at 13.9 nm based on the multilayer beam splitters is used in 13.9 nm soft X-ray laser interference experiment, in which a clear interferograms of C 8 H 8 laser-produced plasma is got. (authors)

  16. A new high quality X-ray source for Cultural Heritage

    Science.gov (United States)

    Walter, Philippe; Variola, Alessandro; Zomer, Fabian; Jaquet, Marie; Loulergue, Alexandre

    2009-09-01

    Compton based photon sources have generated much interest since the rapid advance in laser and accelerator technologies has allowed envisaging their utilisation for ultra-compact radiation sources. These should provide X-ray short pulses with a relatively high average flux. Moreover, the univocal dependence between the scattered photon energy and its angle gives the possibility of obtaining a quasi-monochromatic beam with a simple diaphragm system. For the most ambitious projects the expected performance takes into account a rate of 10-10 photons/s, with an angular divergence of few mrad, an X-ray energy cut-off of few tens of keV and a bandwidth ΔE/E˜1-10%. Even if the integrated rate cannot compete with synchrotron radiation sources, the cost and the compactness of these Compton based machines make them attractive for a wide spectrum of applications. We explore here the interest of these systems for Cultural Heritage preservation. To cite this article: P. Walter et al., C. R. Physique 10 (2009).

  17. Evaluation of bent-crystal x-ray backlighting and microscopy techniques for the Sandia Z machine.

    Science.gov (United States)

    Sinars, Daniel B; Bennett, Guy R; Wenger, David F; Cuneo, Michael E; Porter, John L

    2003-07-01

    X-ray backlighting and microscopy systems for the 1-10-keV range based on spherically or toroidally bent crystals are discussed. These systems are ideal for use on the Sandia Z machine, a megajoule-class x-ray facility. Near-normal-incidence crystal microscopy systems have been shown to be more efficient than pinhole cameras with the same spatial resolution and magnification [Appl. Opt. 37, 1784 (1998)]. We show that high-resolution (< or = 10 microm) x-ray backlighting systems using bent crystals can be more efficient than analogous point-projection imaging systems. Examples of bent-crystal-backlighting results that demonstrate 10-microm resolution over a 20-mm field of view are presented.

  18. Evaluation of bent-crystal x-ray backlighting and microscopy techniques for the Sandia Z machine

    International Nuclear Information System (INIS)

    Sinars, Daniel B.; Wenger, David F.; Cuneo, Michael E.; Porter, John L.; Bennett, Guy R.

    2003-01-01

    X-ray backlighting and microscopy systems for the 1-10-keV range based on spherically or toroidally bent crystals are discussed. These systems are ideal for use on the Sandia Z machine, a megajoule-class x-ray facility. Near-normal-incidence crystal microscopy systems have been shown to be more efficient than pinhole cameras with the same spatial resolution and magnification [Appl. Opt. 37, 1784 (1998)]. We show that high-resolution (≤10 μm) x-ray backlighting systems using bent crystals can be more efficient than analogous point-projection imaging systems. Examples of bent-crystal-backlighting results that demonstrate 10-μm resolution over a 20-mm field of view are presented

  19. DOE/DMS workshop on future synchrotron VUV and x-ray beam Lines

    International Nuclear Information System (INIS)

    Green, P.H.

    1992-03-01

    This document contains an overview of the participating DOE Laboratory beam line interests and the projected science to be addressed on these beam lines, both at new and existing synchrotron facilities. The scientific programs associated with present and planned synchrotron research by DOE Laboratories are discussed in chapters titled ''VUV and Soft X-Ray Research'' and ''Hard X-Ray Research.'' This research encompasses a broad range of the nation's scientific and technical research needs from fundamental to applied, in areas including environmental, biological, and physical sciences; new materials; and energy-related technologies. The projected cost of this proposed construction has been provided in tabular form using a uniform format so that anticipated DOE and outside funding agency contributions for construction and for research and development can be determined. The cost figures are, of course, subject to uncertainties of detailed design requirements and the availability of facility-designed generic components and outside vendors. The report also contains a compendium (as submitted by the beam line proposers) of the design capabilities, the anticipated costs, and the scientific programs of projected beam line construction at the four synchrotron facilities. A summary of the projected cost of these beam lines to be requested of DOE is compiled

  20. Gamma detector for use with luggage X-ray systems

    International Nuclear Information System (INIS)

    McHugh, H.; Quam, W.

    1998-01-01

    A new gamma radiation sensor has been designed for installation on several types of luggage x-ray machines and mobile x-ray vans operated by the U.S. Customs Service and the U.S. Department of State. The use of gamma detectors on x-ray machines imposed difficulties not usually encountered in the design of gamma detectors because the spectrum of scattered x-rays, which varied from machine to machine, extended to energies significantly higher than those of the low-energy isotopic emissions. In the original design, the lower level discriminator was raised above the x-ray end point energy resulting in the loss of the americium line associated with plutonium. This reduced the overall sensitivity to unshielded plutonium by a factor of approximately 100. An improved method was subsequently developed wherein collimation was utilized in conjunction with a variable counting threshold to permit accommodation of differing conditions of x-ray scattering. This design has been shown to eliminate most of the problems due to x-ray scattering while still capturing the americium emissions. The overall sensitivity has remained quite high, though varying slightly from one model of x-ray machine to another, depending upon the x-ray scattering characteristics of each model. (author)

  1. Long-term stability of beam quality and output of conventional X-ray units.

    Science.gov (United States)

    Fukuda, Atsushi; Matsubara, Kosuke; Miyati, Tosiaki

    2015-01-01

    Conventional diagnostic X-ray units are used for radiographic imaging in many countries. For obtaining entrance surface doses, a numerical dose determination method has been applied in Japan. Although this technique is effective, it has to account for errors, particularly fluctuations, due to the beam quality and output of X-ray tubes. As a part of our quality control procedures, we recorded the entrance surface air kerma, tube voltage, and half-value layer measurements made for four diagnostic X-ray tubes over a 103-week period. The entrance surface air kerma for one of the four X-ray tubes had increased significantly by 11.4 % over 1 year from its initial setting, whereas the tube voltages and half-value layers did not deviate significantly from their initial values. Medical physicists and radiological technologists should be aware of this fluctuation for diagnostic X-ray tubes and take it into consideration when calculating the entrance surface air kerma.

  2. A portable Compton spectrometer for clinical X-ray beams in the energy range 20-150 keV

    International Nuclear Information System (INIS)

    Vieira, A.A.; Linke, A.; Yoshimura, E.M.; Terini, R.A.; Herdade, S.B.

    2011-01-01

    Primary beam spectra were obtained for an X-ray industrial equipment (40-150 kV), and for a clinical mammography apparatus (25-35 kV) from beams scattered at angles close to 90 o , measured with a CdTe Compton spectrometer. Actual scattering angles were determined from the Compton energy shift of characteristic X-rays or spectra end-point energy. Evaluated contribution of coherent scattering amounts to more than 15% of fluence in mammographic beams. This technique can be used in clinical environments.

  3. A beam intensity monitor for the evaluation beamline for soft x-ray optical elements

    International Nuclear Information System (INIS)

    Imazono, Takashi; Moriya, Naoji; Harada, Yoshihisa; Sano, Kazuo; Koike, Masato

    2012-01-01

    Evaluation Beamline for Soft X-Ray Optical Elements (BL-11) at the SR Center of Ritsumeikan University has been operated to measure the wavelength and angular characteristics of soft x-ray optical components in a wavelength range of 0.65-25 nm using a reflecto-diffractometer (RD). The beam intensity monitor that has been equipped in BL-11 has observed the signal of the zero-th order light. For the purpose of more accurate evaluation of the performance of optical components, a new beam intensity monitor to measure the intensity of the first order light from the monochromator in BL-11 has been developed and installed in just front of RD. The strong positive correlation between the signal of the beam monitor and a detector equipped in the RD is shown. It is successful that the beam intensity of the first order light can be monitored in real time.

  4. Study of performance of electronic dosemeters in continuous and pulsed X-radiation beams

    International Nuclear Information System (INIS)

    Guimaraes, Margarete Cristina

    2014-01-01

    Personal radiation monitoring is a basic procedure to verify the compliance to regulatory requirements for radiological protection. Electronic personal dosimeters (EPD) based on solid state detectors have largely been used for personnel monitoring; including for pulsed radiation beams where their responses are not well known and deficiencies have been reported. In this work, irradiation conditions for testing the response of EPDs in both continuous and pulsed X-ray beams were studied to be established in a constant potential Seifert-Pantak and in a medical Pulsar 800 Plus VMI X-ray machines. Characterization of X-ray beams was done in terms of tube voltage, half-value layer, mean energy and air kerma rate. A Xi R/F Unfors solid state dosimeter used as reference for air kerma measurements was verified against a RC-6 and 10X6-6 Radical ionization chambers as far its metrological coherence. Rad-60 RADOS, PDM- 11 Aloka and EPD MK2 Thermo electron EPDs were selected to be tested in terms of relative intrinsic error and energy response in similar to IEC RQR, IEC RQA and ISO N reference radiations. Results demonstrated the reliability of the solid state Xi R/F Unfors dosimeter to be as reference dosimeter although its response was affected by heavily filtered beams. Results also showed that relative intrinsic errors in the response of the EPDs in terms of personal dose equivalent, Hp(10), were higher than the requirement established for continuous beams. In pulsed beams, some EPDs showed inadequate response and high relative intrinsic errors. This work stressed the need of performing additional checks for EPDs, besides the limited 137 Cs beam calibration, before using them in pulsed X-ray beams. (author)

  5. Shielding requirements for constant-potential diagnostic x-ray beams determined by a Monte Carlo calculation

    International Nuclear Information System (INIS)

    Simpkin, D.J.

    1989-01-01

    A Monte Carlo calculation has been performed to determine the transmission of broad constant-potential x-ray beams through Pb, concrete, gypsum wallboard, steel and plate glass. The EGS4 code system was used with a simple broad-beam geometric model to generate exposure transmission curves for published 70, 100, 120 and 140-kVcp x-ray spectra. These curves are compared to measured three-phase generated x-ray transmission data in the literature and found to be reasonable. For calculation ease the data are fit to an equation previously shown to describe such curves quite well. These calculated transmission data are then used to create three-phase shielding tables for Pb and concrete, as well as other materials not available in Report No. 49 of the NCRP

  6. Shielding requirements for constant-potential diagnostic x-ray beams determined by a Monte Carlo calculation.

    Science.gov (United States)

    Simpkin, D J

    1989-02-01

    A Monte Carlo calculation has been performed to determine the transmission of broad constant-potential x-ray beams through Pb, concrete, gypsum wallboard, steel and plate glass. The EGS4 code system was used with a simple broad-beam geometric model to generate exposure transmission curves for published 70, 100, 120 and 140-kVcp x-ray spectra. These curves are compared to measured three-phase generated x-ray transmission data in the literature and found to be reasonable. For calculation ease the data are fit to an equation previously shown to describe such curves quite well. These calculated transmission data are then used to create three-phase shielding tables for Pb and concrete, as well as other materials not available in Report No. 49 of the NCRP.

  7. Development of a PIXE (Particle Induced X-ray Emission) analysis device using an extracted proton beam

    International Nuclear Information System (INIS)

    Saidi, A.

    1989-01-01

    The experimental device described allows the extention of the PIXE (Particle Induced X-ray Emission) method to the analysis, by means of proton beams, of solid or liquid samples, which can not be analyzed under vacuum conditions. The homogeneity of the surfaces to be analysed and elements (in the atmosphere) which absorb X-rays must be taken into account. Liquid samples do not need special care. The results show that: at high energies, the extracted beam sensibility is of the same order of magnitude as those obtained under vacuum; at low energies, the performance under vacuum conditions is better. The particles energy losses, at the exit membrane and in the outer atmosphere, decrease the X-rays production efficiency [fr

  8. Soft X-ray radiation damage in EM-CCDs used for Resonant Inelastic X-ray Scattering

    Science.gov (United States)

    Gopinath, D.; Soman, M.; Holland, A.; Keelan, J.; Hall, D.; Holland, K.; Colebrook, D.

    2018-02-01

    Advancement in synchrotron and free electron laser facilities means that X-ray beams with higher intensity than ever before are being created. The high brilliance of the X-ray beam, as well as the ability to use a range of X-ray energies, means that they can be used in a wide range of applications. One such application is Resonant Inelastic X-ray Scattering (RIXS). RIXS uses the intense and tuneable X-ray beams in order to investigate the electronic structure of materials. The photons are focused onto a sample material and the scattered X-ray beam is diffracted off a high resolution grating to disperse the X-ray energies onto a position sensitive detector. Whilst several factors affect the total system energy resolution, the performance of RIXS experiments can be limited by the spatial resolution of the detector used. Electron-Multiplying CCDs (EM-CCDs) at high gain in combination with centroiding of the photon charge cloud across several detector pixels can lead to sub-pixel spatial resolution of 2-3 μm. X-ray radiation can cause damage to CCDs through ionisation damage resulting in increases in dark current and/or a shift in flat band voltage. Understanding the effect of radiation damage on EM-CCDs is important in order to predict lifetime as well as the change in performance over time. Two CCD-97s were taken to PTB at BESSY II and irradiated with large doses of soft X-rays in order to probe the front and back surfaces of the device. The dark current was shown to decay over time with two different exponential components to it. This paper will discuss the use of EM-CCDs for readout of RIXS spectrometers, and limitations on spatial resolution, together with any limitations on instrument use which may arise from X-ray-induced radiation damage.

  9. Reproducibility of temporomandibular joint tomography. Influence of shifted X-ray beam and tomographic focal plane on reproducibility

    International Nuclear Information System (INIS)

    Saito, Masashi

    1999-01-01

    Proper tomographic focal plane and x-ray beam direction are the most important factors to obtain accurate images of the temporomandibular joint (TMJ). In this study, to clarify the magnitude of effect of these two factors on the image quality. We evaluated the reproducibility of tomograms by measuring the distortion when the x-ray beam was shifted from the correct center of the object. The effects of the deviation of the tomographic focal plane on image quality were evaluated by the MTF (Modulation Transfer Function). Two types of tomograms, one the plane type, the other the rotational type were used in this study. A TMJ model was made from Teflon for the purpose of evaluation by shifting the x-ray beam. The x-ray images were obtained by tilting the model from 0 to 10 degrees 2-degree increments. These x-ray images were processed for computer image analysis, and then the distance between condyle and the joint space was measured. To evaluate the influence of the shifted tomographic focal plane on image sharpness, the x-ray images from each setting were analyzed by MTF. To obtain the MTF, ''knife-edge'' made from Pb was used. The images were scanned with a microdensitometer at the central focal plane, and 0, 0.5, 1 mm away respectively. The density curves were analyzed by Fourier analysis and the MTF was calculated. The reproducibility of images became worse by shifting the x-ray beam. This tendency was similar for both tomograms. Object characteristics such as anterior and posterior portion of the joint space affected the deterioration of reproducibility of the tomography. The deviation of the tomographic focal plane also decreased the reproducibility of the x-ray images. The rotational type showed a better MTF, but it became seriously unfavorable with slight changes of the tomographic focal plane. Contrarily, the plane type showed a lower MTF, but the image was stable with shifting of the tomographic focal plane. (author)

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that ...

  11. Stabilized x-ray generator power supply

    International Nuclear Information System (INIS)

    Saha, Subimal; Purushotham, K.V.; Bose, S.K.

    1986-01-01

    X-ray diffraction and X-ray fluorescence analysis are very much adopted in laboratories to determine the type and structure of the constituent compounds in solid materials, chemical composition of materials, stress developed on metals etc. These experiments need X-ray beam of fixed intensity and wave length. This can only be achieved by X-ray generator having highly stabilized tube voltage and tube current. This paper describes how X-ray tube high voltage and electron beam current are stabilized. This paper also highlights generation of X-rays, diffractometry and X-ray fluorescence analysis and their wide applications. Principle of operation for stabilizing the X-ray tube voltage and current, different protection circuits adopted, special features of the mains H.V. transformer and H.T. tank are described in this report. (author)

  12. Generalized eMC implementation for Monte Carlo dose calculation of electron beams from different machine types.

    Science.gov (United States)

    Fix, Michael K; Cygler, Joanna; Frei, Daniel; Volken, Werner; Neuenschwander, Hans; Born, Ernst J; Manser, Peter

    2013-05-07

    The electron Monte Carlo (eMC) dose calculation algorithm available in the Eclipse treatment planning system (Varian Medical Systems) is based on the macro MC method and uses a beam model applicable to Varian linear accelerators. This leads to limitations in accuracy if eMC is applied to non-Varian machines. In this work eMC is generalized to also allow accurate dose calculations for electron beams from Elekta and Siemens accelerators. First, changes made in the previous study to use eMC for low electron beam energies of Varian accelerators are applied. Then, a generalized beam model is developed using a main electron source and a main photon source representing electrons and photons from the scattering foil, respectively, an edge source of electrons, a transmission source of photons and a line source of electrons and photons representing the particles from the scrapers or inserts and head scatter radiation. Regarding the macro MC dose calculation algorithm, the transport code of the secondary particles is improved. The macro MC dose calculations are validated with corresponding dose calculations using EGSnrc in homogeneous and inhomogeneous phantoms. The validation of the generalized eMC is carried out by comparing calculated and measured dose distributions in water for Varian, Elekta and Siemens machines for a variety of beam energies, applicator sizes and SSDs. The comparisons are performed in units of cGy per MU. Overall, a general agreement between calculated and measured dose distributions for all machine types and all combinations of parameters investigated is found to be within 2% or 2 mm. The results of the dose comparisons suggest that the generalized eMC is now suitable to calculate dose distributions for Varian, Elekta and Siemens linear accelerators with sufficient accuracy in the range of the investigated combinations of beam energies, applicator sizes and SSDs.

  13. Long-term stability of beam quality and output of conventional X-ray units

    OpenAIRE

    Fukuda, Atsushi; Matsubara, Kosuke; Miyati, Toshiaki

    2015-01-01

    Conventional diagnostic X-ray units are used for radiographic imaging in many countries. For obtaining entrance surface doses, a numerical dose determination method has been applied in Japan. Although this technique is effective, it has to account for errors, particularly fluctuations, due to the beam quality and output of X-ray tubes. As a part of our quality control procedures, we recorded the entrance surface air kerma, tube voltage, and half-value layer measurements made for four diagnost...

  14. Advances in thin film diffraction instrumentation by X-ray optics

    International Nuclear Information System (INIS)

    Haase, A.

    1996-01-01

    The structural characterisation of thin films requires a parallel X-ray beam of high intensity. Parallel beam geometry is commonly used in high resolution and single crystal experiments, but also in the field of X-ray diffraction for polycrystalline material (e.g. in phase, texture and stress analysis). For grazing incidence diffraction (GID), the use of small slits on the primary side and of long soller slits with a flat monochromator on the secondary side is standard. New optical elements have been introduced with polychromatic or monochromatic radiation. By means of different applications the results are compared with those of classical beam optics. X-ray fiber optics utilize total external reflection of X-rays on smooth surfaces. Effects of monochromatization are presented. In many fields of application, fiber optics may replace conventional collimators. The use of primary and secondary channel cut crystals can also produce a high parallel monochromatic X-ray beam. A parabolically bent graded multilayer produces a monochromatic parallel beam of high intensity. Compared with classical Bragg-Brentano (focussing) geometry, excellent results have been obtained, especially for samples with an irregular shape. In combination with a channel cut monochromator there is a substantial gain in intensity leading to an increase of the dynamic intensity range of rocking curves

  15. Advances in thin film diffraction instrumentation by X-ray optics

    Energy Technology Data Exchange (ETDEWEB)

    Haase, A [Rich. Seifert and Co., Analytical X-ray Systems, Ahrensburg (Germany)

    1996-09-01

    The structural characterisation of thin films requires a parallel X-ray beam of high intensity. Parallel beam geometry is commonly used in high resolution and single crystal experiments, but also in the field of X-ray diffraction for polycrystalline material (e.g. in phase, texture and stress analysis). For grazing incidence diffraction (GID), the use of small slits on the primary side and of long soller slits with a flat monochromator on the secondary side is standard. New optical elements have been introduced with polychromatic or monochromatic radiation. By means of different applications the results are compared with those of classical beam optics. X-ray fiber optics utilize total external reflection of X-rays on smooth surfaces. Effects of monochromatization are presented. In many fields of application, fiber optics may replace conventional collimators. The use of primary and secondary channel cut crystals can also produce a high parallel monochromatic X-ray beam. A parabolically bent graded multilayer produces a monochromatic parallel beam of high intensity. Compared with classical Bragg-Brentano (focussing) geometry, excellent results have been obtained, especially for samples with an irregular shape. In combination with a channel cut monochromator there is a substantial gain in intensity leading to an increase of the dynamic intensity range of rocking curves.

  16. Nanofocusing refractive X-ray lenses

    Energy Technology Data Exchange (ETDEWEB)

    Boye, Pit

    2010-02-05

    This thesis is concerned with the optimization and development of the production of nanofocusing refractive X-ray lenses. These optics made of either silicon or diamond are well-suited for high resolution X-ray microscopy. The goal of this work is the design of a reproducible manufacturing process which allows the production of silicon lenses with high precision, high quality and high piece number. Furthermore a process for the production of diamond lenses is to be developed and established. In this work, the theoretical basics of X-rays and their interaction with matter are described. Especially, aspects of synchrotron radiation are emphasized. Important in X-ray microscopy are the different optics. The details, advantages and disadvantages, in particular those of refractive lenses are given. To achieve small X-ray beams well beyond the 100 nm range a small focal length is required. This is achieved in refractive lenses by moving to a compact lens design where several single lenses are stacked behind each other. The, so-called nanofocusing refractive lenses (NFLs) have a parabolic cylindrical shape with lateral structure sizes in the micrometer range. NFLs are produced by using micro-machining techniques. These micro-fabrication processes and technologies are introduced. The results of the optimization and the final fabrication process for silicon lenses are presented. Subsequently, two experiments that are exemplary for the use of NFLs, are introduced. The rst one employs a high-resolution scanning fluorescence mapping of a geological sample, and the second one is a coherent x-ray diffraction imaging (CXDI) experiment. CXDI is able to reconstruct the illuminated object from recorded coherent diffraction patterns. In a scanning mode, referred to as ptychography, this method is even able to reconstruct the illumination and the object simultaneously. Especially the reconstructed illumination and the possibility of computed propagation of the wave field along the

  17. Nanofocusing refractive X-ray lenses

    International Nuclear Information System (INIS)

    Boye, Pit

    2010-01-01

    This thesis is concerned with the optimization and development of the production of nanofocusing refractive X-ray lenses. These optics made of either silicon or diamond are well-suited for high resolution X-ray microscopy. The goal of this work is the design of a reproducible manufacturing process which allows the production of silicon lenses with high precision, high quality and high piece number. Furthermore a process for the production of diamond lenses is to be developed and established. In this work, the theoretical basics of X-rays and their interaction with matter are described. Especially, aspects of synchrotron radiation are emphasized. Important in X-ray microscopy are the different optics. The details, advantages and disadvantages, in particular those of refractive lenses are given. To achieve small X-ray beams well beyond the 100 nm range a small focal length is required. This is achieved in refractive lenses by moving to a compact lens design where several single lenses are stacked behind each other. The, so-called nanofocusing refractive lenses (NFLs) have a parabolic cylindrical shape with lateral structure sizes in the micrometer range. NFLs are produced by using micro-machining techniques. These micro-fabrication processes and technologies are introduced. The results of the optimization and the final fabrication process for silicon lenses are presented. Subsequently, two experiments that are exemplary for the use of NFLs, are introduced. The rst one employs a high-resolution scanning fluorescence mapping of a geological sample, and the second one is a coherent x-ray diffraction imaging (CXDI) experiment. CXDI is able to reconstruct the illuminated object from recorded coherent diffraction patterns. In a scanning mode, referred to as ptychography, this method is even able to reconstruct the illumination and the object simultaneously. Especially the reconstructed illumination and the possibility of computed propagation of the wave field along the

  18. Effect of X-ray suppression system upon parameters of electrostatic accelerator ion beam

    Directory of Open Access Journals (Sweden)

    I. G. Ignat'ev

    2014-12-01

    Full Text Available Experimental study results are presented for a beam profile and emittance of an electrostatic accelerator “Sokol” before and after being equipped with magnet X-ray suppression system.

  19. X-ray section topographs under various coherence properties of the primary beam

    International Nuclear Information System (INIS)

    Borowski, J.; Gronkowski, J.

    2001-01-01

    The aim of this work is to study to what extent a typical section-topography setup can supply information about the degree of coherence of the incident x-ray beam. In real experiments, the incident beam is partially coherent, with the degree of coherence described by the shape of the correlation function. In this paper the correlation functions for the outgoing beam are calculated by solving the Takagi-Taupin equations, assuming a truncated Gauss correlation function for the incident beam with the correlation length determined by the van Cittert-Zernike theorem. Its influence on the measured intensity of the diffracted beam in section topography is investigated. (author)

  20. Update on VUV and soft X-ray facilities at SSRL

    International Nuclear Information System (INIS)

    Waldhauer, A.

    1988-01-01

    The number of experimental stations at SSRL devoted to the VUV and soft X-ray region is increasing rapidly. In 1986 there were five VUV/soft X-ray beam lines in regular operation. These consisted of two grasshopper lines, a Seya-Namioka line, a white light lithography line, and the UHV double crystal line, Jumbo. By 1988 ten beam lines, including two with insertion devices, covering the spectral range 5-4000 eV in five overlapping ranges will be operational. With the addition of these new stations, SSRL will have increased dramatically its facilities for performing VUV and soft X-ray research. (orig.)

  1. X-ray diffraction imaging of material microstructures

    KAUST Repository

    Varga, Laszlo

    2016-10-20

    Various examples are provided for x-ray imaging of the microstructure of materials. In one example, a system for non-destructive material testing includes an x-ray source configured to generate a beam spot on a test item; a grid detector configured to receive x- rays diffracted from the test object; and a computing device configured to determine a microstructure image based at least in part upon a diffraction pattern of the x-rays diffracted from the test object. In another example, a method for determining a microstructure of a material includes illuminating a beam spot on the material with a beam of incident x-rays; detecting, with a grid detector, x-rays diffracted from the material; and determining, by a computing device, a microstructure image based at least in part upon a diffraction pattern of the x-rays diffracted from the material.

  2. Numerical controlled diamond fly cutting machine for grazing incidence X-ray reflection mirrors

    International Nuclear Information System (INIS)

    Uchida, Fumihiko; Moriyama, Shigeo; Seya, Eiiti

    1992-01-01

    Synchrotron radiation has reached the stage of practical use, and the application to the wide fields that support future advanced technologies such as spectroscopy, the structural analysis of matters, semiconductor lithography and medical light source is expected. For the optical system of the equipment utilizing synchrotron radiation, the total reflection mirrors of oblique incidence are used for collimating and collecting X-ray. In order to restrain their optical aberration, nonspherical shape is required, and as the manufacturing method with high precision for nonspherical mirrors, a numerically controlled diamond cutting machine was developed. As for the cutting of soft metals with diamond tools, the high precision machining of any form can be done by numerical control, the machining time can be reduced as compared with grinding, and the cooling effect is large in metals. The construction of the cutting machine, the principle of machining, the control system, the method of calculating numerical control data, the investigation of machinable forms and the result of evaluation are reported. (K.I.)

  3. Spectroscopic analysis and dosimetry of diagnostic x-ray beams filtered by rare earth materials

    International Nuclear Information System (INIS)

    Tyndall, D.A.

    1986-01-01

    A laboratory investigation was carried out to assess the effect of various types of rare earth filter materials on the energy spectrum and concomitant reduced exposure values of diagnostic x-ray beams at 70, 80, and 90 kVp. An x-ray spectroscope was constructed and used to generate the energy spectra of beams passing through the various rare earth filter materials. Photographs were made of each spectrum, and live-time gross photon counts were recorded. Following spectral determinations, ionization chamber readings were generated for each filter material. Substantial effects on x-ray spectra and reduction of exposure values were noted. The degree of these effects were dependent on the atomic number, k-edge, and thickness of each filter. Metallic forms of rare earth materials proved to be more effective than the salt forms with erbium offering the greatest potential for reduction in exposures over the range of experimental kilovolt (peak) values

  4. X-Ray Beam Studies of Charge Sharing in Small Pixel, Spectroscopic, CdZnTe Detectors

    Science.gov (United States)

    Allwork, Christopher; Kitou, Dimitris; Chaudhuri, Sandeep; Sellin, Paul J.; Seller, Paul; Veale, Matthew C.; Tartoni, Nicola; Veeramani, Perumal

    2012-08-01

    Recent advances in the growth of CdZnTe material have allowed the development of small pixel, spectroscopic, X-ray imaging detectors. These detectors have applications in a diverse range of fields such as medical, security and industrial sectors. As the size of the pixels decreases relative to the detector thickness, the probability that charge is shared between multiple pixels increases due to the non zero width of the charge clouds drifting through the detector. These charge sharing events will result in a degradation of the spectroscopic performance of detectors and must be considered when analyzing the detector response. In this paper charge sharing and charge loss in a 250 μm pitch CdZnTe pixel detector has been investigated using a mono-chromatic X-ray beam at the Diamond Light Source, U.K. Using a 20 μm beam diameter the detector response has been mapped for X-ray energies both above (40 keV) and below (26 keV) the material K-shell absorption energies to study charge sharing and the role of fluorescence X-rays in these events.

  5. A Monte Carlo approach for simulating the propagation of partially coherent x-ray beams

    DEFF Research Database (Denmark)

    Prodi, A.; Bergbäck Knudsen, Erik; Willendrup, Peter Kjær

    2011-01-01

    Advances at SR sources in the generation of nanofocused beams with a high degree of transverse coherence call for effective techniques to simulate the propagation of partially coherent X-ray beams through complex optical systems in order to characterize how coherence properties such as the mutual...

  6. Ultra-short wavelength x-ray system

    Science.gov (United States)

    Umstadter, Donald [Ann Arbor, MI; He, Fei [Ann Arbor, MI; Lau, Yue-Ying [Potomac, MD

    2008-01-22

    A method and apparatus to generate a beam of coherent light including x-rays or XUV by colliding a high-intensity laser pulse with an electron beam that is accelerated by a synchronized laser pulse. Applications include x-ray and EUV lithography, protein structural analysis, plasma diagnostics, x-ray diffraction, crack analysis, non-destructive testing, surface science and ultrafast science.

  7. SU-E-T-143: Effect of X-Ray and Cone Beam CT Reconstruction Parameters On Estimation of Bone Volume of Mice Used in Aging Research

    Energy Technology Data Exchange (ETDEWEB)

    Russ, M; Pang, M; Troen, B; Rudin, S; Ionita, C [University at Buffalo, Buffalo, NY (United States)

    2014-06-01

    Purpose: To investigate the variations in bone volume calculations in mice involved in aging research when changing cone beam micro-CT x-ray and reconstruction parameters. Methods: Mouse spines were placed on an indexed turn table that rotated 0.5° per projection and imaged by a self-built micro CT machine containing a CCD-based high-resolution x-ray detector. After the full 360° rotation data set of object images was obtained, a standard filtered back-projection cone beam reconstruction was performed. Four different kVp's between 40–70 kVp in 10kVp increments were selected. For each kVp two mAs settings were used. Each acquisition was reconstructed using two voxel sizes (12 and 25μm) and two step angles, 0.5° and 1°, respectively. A LabView program was written to determine the total bone volume contained in the mouse's total spine volume (bone plus gaps) as a measure of spine health. First, the user selected the desired 512×512 reconstruction to view the whole spine volume which was then used to select a gray-level threshold that allowed for viewing of the bone structure, then another threshold to include gaps. The program returned bone volume, bone × gap volume, and their ratio, BVF. Results: The calculated bone volume fractions were compared as a function of tube potential. Cases with 25μm slice thickness showed trials with lower kVp's had greater image contrast, which resulted in higher calculated bone volume fractions. Cases with 12μm reconstructed slice thickness were significantly noisier, and showed no clear maximum BVF. Conclusion: Using the projection images and reconstructions acquired from the micro CT, it can be shown that the micro-CT x-ray and reconstruction parameters significantly affect the total bone volume calculations. When comparing mice cohorts treated with different therapies researchers need to be aware of such details and use volumes which were acquired and processed in identical conditions.

  8. Generalized algorithm for X-ray projections generation in cone-beam tomography

    International Nuclear Information System (INIS)

    Qin Zhongyuan; Mu Xuanqin; Wang Ping; Cai Yuanlong; Hou Chuanjian

    2002-01-01

    In order to get rid of random factors in the measurement so as to support proceeding 3D reconstruction, a general approach is presented to obtain the X-ray projections in cone-beam tomography. The phantom is firstly discretized into cubic volume through inverse transformation then a generalized projection procedure is proposed to the digitized result without concerning what the phantom exactly is. In the second step, line integrals are calculated to obtain the projection of each X-ray through accumulation of tri-linear interpolation. Considering projection angles, a rotation matrix is proposed to the X-ray source and the detector plane so projections in arbitrary angles can be got. In this approach the algorithm is easy to be extended and irregular objects can also be processed. The algorithm is implemented in Visual C++ and experiments are done using different models. Satisfactory results are obtained. It makes good preparation for the proceeding reconstruction

  9. Hard alpha-keratin degradation inside a tissue under high flux X-ray synchrotron micro-beam: a multi-scale time-resolved study.

    Science.gov (United States)

    Leccia, Emilie; Gourrier, Aurélien; Doucet, Jean; Briki, Fatma

    2010-04-01

    X-rays interact strongly with biological organisms. Synchrotron radiation sources deliver very intense X-ray photon fluxes within micro- or submicro cross-section beams, resulting in doses larger than the MGy. The relevance of synchrotron radiation analyses of biological materials is therefore questionable since such doses, million times higher than the ones used in radiotherapy, can cause huge damages in tissues, with regard to not only DNA, but also proteic and lipid organizations. Very few data concerning the effect of very high X-ray doses in tissues are available in the literature. We present here an analysis of the structural phenomena which occur when the model tissue of human hair is irradiated by a synchrotron X-ray micro-beam. The choice of hair is supported by its hierarchical and partially ordered keratin structure which can be analysed inside the tissue by X-ray diffraction. To assess the damages caused by hard X-ray micro-beams (1 microm(2) cross-section), short exposure time scattering SAXS/WAXS patterns have been recorded at beamline ID13 (ESRF) after various irradiation times. Various modifications of the scattering patterns are observed, they provide fine insight of the radiation damages at various hierarchical levels and also unexpectedly provide information about the stability of the various hierarchical structural levels. It appears that the molecular level, i.e. the alpha helices which are stabilized by hydrogen bonds and the alpha-helical coiled coils which are stabilized by hydrophobic interactions, is more sensitive to radiation than the supramolecular architecture of the keratin filament and the filament packing within the keratin associated proteins matrix, which is stabilized by disulphide bonds. (c) 2009 Elsevier Inc. All rights reserved.

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... evaluation. National and international radiology protection organizations continually review and update the technique standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  11. Coherent hard x-ray focusing optics and applications

    Energy Technology Data Exchange (ETDEWEB)

    Yun, W.B.; Viccaro, P.J.; Chrzas, J.; Lai, B.

    1991-01-01

    Coherent hard x-ray beams with a flux exceeding 10{sup 9} photons/second with a bandwidth of 0.1% will be provided by the undulator at the third generation synchrotron radiation sources such as APS, ESRF, and Spring-8. The availability of such high flux coherent x-ray beams offers excellent opportunities for extending the coherence-based techniques developed in the visible and soft x-ray part of the electromagnetic spectrum to the hard x-rays. These x-ray techniques (e.g., diffraction limited microfocusing, holography, interferometry, phase contrast imaging and signal enhancement), may offer substantial advantages over non-coherence-based x-ray techniques currently used. For example, the signal enhancement technique may be used to enhance an anomalous x-ray or magnetic x-ray scattering signal by several orders of magnitude. Coherent x-rays can be focused to a very small (diffraction-limited) spot size, thus allowing high spatial resolution microprobes to be constructed. The paper will discuss the feasibility of the extension of some coherence-based techniques to the hard x-ray range and the significant progress that has been made in the development of diffraction-limited focusing optics. Specific experimental results for a transmission Fresnel phase zone plate that can focus 8.2 keV x-rays to a spot size of about 2 microns will be briefly discussed. The comparison of measured focusing efficiency of the zone plate with that calculated will be made. Some specific applications of zone plates as coherent x-ray optics will be discussed. 17 refs., 4 figs.

  12. Machine learning from hard x-ray surveys: applications to magnetic cataclysmic variable studies

    Science.gov (United States)

    Scaringi, Simone

    2009-11-01

    Within this thesis are discussed two main topics of contemporary astrophysics. The first is that of machine learning algorithms for astronomy whilst the second is that of magnetic cataclysmic variables (mCVs). To begin, an overview is given of ISINA: INTEGRAL Scouce Identifiction Network Algorithm. This machine learning algorithm, using random forests, is applied to the IBIS/ISGRI data set in order to ease the production of unbiased future soft gamma-ray source catalogues. The feature extraction process on an initial candidate list is described together with feature merging. Three trainng and testing sets are created in order to deal with the diverse time-scales encountered when dealing with the gamma-ray sky: one dealing with faint persistent source recognition, one dealing with strong persistent sources and a final one dealing with transients. For the latter, a new transient detection technique is introduced and described: the transient matrix. Finally the performance of the network is assessed and discussed using the testing set and some illustrative source examples. ISINA is also compared to the more conventional approach of visual inspection. Next mCVs are discussed, and in particular the properties arising from a hard X-ray selected sample which has proven remarkably efficient in detecting intermediate polars and asynchronous polars, two of the rarest type of cataclysmic variables (CVs). This thesis focuses particularly on the link between hard X-ray properties and spin/orbital periods. To this end, a new sample of these objects is constructed by cross-corelating candidate sources detected in INTEGRAL/IBIS observations against catalogues of known CVs. Also included in the analysis are hard X-ray Observations from Swift/BAT and SUZAKU/HXD in order to make the study more complete. It is found that most hard X-ray detected mCVs have Pspin/Porb<0.1 above the period gap. In this respect, attention is given to the very low number of detected systems in any ban

  13. X-ray beam penetration in TXRF measurement of polycrystalline and amorphous surfaces

    International Nuclear Information System (INIS)

    Ghatak-Roy, A.R.; Hossain, T.Z.

    2000-01-01

    For TXRF measurement on single crystal silicon surface, it is generally agreed that the x-ray beam penetration is of the order of a few hundred Angstroms from the surface. However, for polycrystalline and amorphous surfaces - frequently used in semiconductor manufacturing there are evidences that x-rays penetrate much deeper revealing underlying layers. The evidences come from various measurements done with films such as silicon dioxide, silicon nitride and metal films such as aluminum, titanium and cobalt. A systematic study was carried out to help understand the issue further. Four sets of samples (on 8 inch wafers) were prepared to create layers buried under various deposited metal and non-metal layers. The metal layers created were aluminum, titanium and cobalt and the non-metal layers were silicon dioxide and silicon nitride. These samples were analyzed by TXRF under various angles and energies and the data were analyzed for signals from various buried layers along with their angular dependence. The results indicated deep penetration of x-ray beams. The samples were further analyzed by SIMS (Secondary Ion Mass Spectrometry) and some of them by ICP-MS (Inductively Coupled Plasma - Mass Spectrometry) to obtain information about their depth profiles. This was done in order to rule out the possibility of intermixing of layers during deposition. (author)

  14. Fluorescence detection of white-beam X-ray absorption anisotropy: towards element-sensitive projections of local atomic structure

    International Nuclear Information System (INIS)

    Korecki, P.; Tolkiehn, M.; Dąbrowski, K. M.; Novikov, D. V.

    2011-01-01

    A method for a direct measurement of X-ray projections of the atomic structure is described. Projections of the atomic structure around Nb atoms in a LiNbO 3 single crystal were obtained from a white-beam X-ray absorption anisotropy pattern detected using Nb K fluorescence. Projections of the atomic structure around Nb atoms in a LiNbO 3 single crystal were obtained from a white-beam X-ray absorption anisotropy (XAA) pattern detected using Nb K fluorescence. This kind of anisotropy results from the interference of X-rays inside a sample and, owing to the short coherence length of a white beam, is visible only at small angles around interatomic directions. Consequently, the main features of the recorded XAA corresponded to distorted real-space projections of dense-packed atomic planes and atomic rows. A quantitative analysis of XAA was carried out using a wavelet transform and allowed well resolved projections of Nb atoms to be obtained up to distances of 10 Å. The signal of nearest O atoms was detected indirectly by a comparison with model calculations. The measurement of white-beam XAA using characteristic radiation indicates the possibility of obtaining element-sensitive projections of the local atomic structure in more complex samples

  15. Characterization of low energy X-rays beams with an extrapolation chamber

    International Nuclear Information System (INIS)

    Bastos, Fernanda Martins

    2015-01-01

    In laboratories involving Radiological Protection practices, it is usual to use reference radiations for calibrating dosimeters and to study their response in terms of energy dependence. The International Organization for Standardization (ISO) established four series of reference X-rays beams in the ISO- 4037 standard: the L and H series, as low and high air Kerma rates, respectively, the N series of narrow spectrum and W series of wide spectrum. The X-rays beams with tube potential below 30 kV, called 'low energy beams' are, in most cases, critical as far as the determination of their parameters for characterization purpose, such as half-value layer. Extrapolation chambers are parallel plate ionization chambers that have one mobile electrode that allows variation of the air volume in its interior. These detectors are commonly used to measure the quantity Absorbed Dose, mostly in the medium surface, based on the extrapolation of the linear ionization current as a function of the distance between the electrodes. In this work, a characterization of a model 23392 PTW extrapolation chamber was done in low energy X-rays beams of the ISO- 4037 standard, by determining the polarization voltage range through the saturation curves and the value of the true null electrode spacing. In addition, the metrological reliability of the extrapolation chamber was studied with measurements of the value of leakage current and repeatability tests; limit values were established for the proper use of the chamber. The PTW23392 extrapolation chamber was calibrated in terms of air Kerma in some of the ISO radiation series of low energy; the traceability of the chamber to the National Standard Dosimeter was established. The study of energy dependency of the extrapolation chamber and the assessment of the uncertainties related to the calibration coefficient were also done; it was shown that the energy dependence was reduced to 4% when the extrapolation technique was used. Finally, the first

  16. Development of high-resolution x-ray CT system using parallel beam geometry

    Energy Technology Data Exchange (ETDEWEB)

    Yoneyama, Akio, E-mail: akio.yoneyama.bu@hitachi.com; Baba, Rika [Central Research Laboratory, Hitachi Ltd., Hatoyama, Saitama (Japan); Hyodo, Kazuyuki [Institute of Materials Science, High Energy Accelerator Research Organization, Tsukuba, Ibaraki (Japan); Takeda, Tohoru [School of Allied Health Sciences, Kitasato University, Sagamihara, Kanagawa (Japan); Nakano, Haruhisa; Maki, Koutaro [Department of Orthodontics, School of Dentistry Showa University, Ota-ku, Tokyo (Japan); Sumitani, Kazushi; Hirai, Yasuharu [Kyushu Synchrotron Light Research Center, Tosu, Saga (Japan)

    2016-01-28

    For fine three-dimensional observations of large biomedical and organic material samples, we developed a high-resolution X-ray CT system. The system consists of a sample positioner, a 5-μm scintillator, microscopy lenses, and a water-cooled sCMOS detector. Parallel beam geometry was adopted to attain a field of view of a few mm square. A fine three-dimensional image of birch branch was obtained using a 9-keV X-ray at BL16XU of SPring-8 in Japan. The spatial resolution estimated from the line profile of a sectional image was about 3 μm.

  17. X-ray Radiation Mechanisms and the Beaming Effect of Hot Spots ...

    Indian Academy of Sciences (India)

    Astr. (2011) 32, 193–196 c Indian Academy of Sciences. X-ray Radiation Mechanisms and the Beaming Effect of Hot Spots and Knots in AGN Jets. Jin Zhang1,∗. , Jin-Ming Bai2, Liang Chen2 & Enwei Liang3. 1College of Physics and Electronic Engineering, Guangxi Teachers Education University,. Nanning 530001, China.

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... are a form of radiation like light or radio waves. X-rays pass through most objects, including the body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small burst of radiation that passes through the body, recording ...

  19. Thermometric- and Acoustic-Based Beam Power Monitor for Ultra-Bright X-Rays

    International Nuclear Information System (INIS)

    2010-01-01

    A design for an average beam power monitor for ultra-bright X-ray sources is proposed that makes simultaneous use of calorimetry and radiation acoustics. Radiation incident on a solid target will induce heating and ultrasonic vibrations, both of which may be measured to give a fairly precise value of the beam power. The monitor is intended for measuring ultra-bright Free-Electron Laser (FEL) X-ray beams, for which traditional monitoring technologies such as photo-diodes or scintillators are unsuitable. The monitor consists of a Boron Carbide (B 4 C) target designed to absorb most of the incident beam's energy. Resistance temperature detectors (RTD) and piezoelectric actuators are mounted on the outward faces of the target to measure the temperature changes and ultrasonic vibrations induced by the incident beam. The design was tested using an optical pulsed beam (780 nm, 120 and 360 Hz) from a Ti:sapphire oscillator at several energies between 0.8 and 2.6 mJ. The RTDs measured an increase in temperature of about 10 K over a period of several minutes. The piezoelectric sensors recorded ringing acoustic oscillations at 580 ± 40 kHz. Most importantly, the amplitude of the acoustic signals was observed to scale linearly with beam power up to 2 mJ of pulse energy. Above this pulse energy, the vibrational signals became nonlinear. Several causes for this nonlinearity are discussed, including amplifier saturation and piezoelectric saturation. Despite this nonlinearity, these measurements demonstrate the feasibility of such a beam power measurement device. The advantage of two distinct measurements (acoustic and thermometric) provides a useful method of calibration that is unavailable to current LCLS diagnostics tools.

  20. Tire inspection machine presenting an X-ray image of the entire width of the tire

    International Nuclear Information System (INIS)

    1975-01-01

    An X-ray machine for production line inspection of tires places each tire in turn with one sector of the tire inside of a U-shaped arrangement of three fluorescent screens so that one screen faces the tread and the other two face the sides of the tire. An X-ray source inside the bead circle of the tire produces X-ray shadow pictures on the screens, which are separately viewed by a video camera, with an image of the picture on each screen transferred to a third of the target of a storage tube, so that a composite image of the entire width of the tire sector can be displayed on a picture tube. The tire is rotated step-wise for display of still pictures of every sector of the tire. (Auth.)

  1. Determination of dose enhancement caused by gold-nanoparticles irradiated with proton, X-rays (kV and MV) and electron beams, using alanine/EPR dosimeters

    International Nuclear Information System (INIS)

    Smith, Clare L.; Ackerly, Trevor; Best, Stephen P.; Gagliardi, Frank; Kie, Katahira; Little, Peter J.; McCorkell, Giulia; Sale, Charlotte A.; Tsunei, Yusuke; Tominaga, Takahiro; Volaric, Sioe See; Geso, Moshi

    2015-01-01

    The main aims of this research was to employ alanine doped with gold-nanoparticles “AuNPs” to determine the levels of dose enhancement caused by these particles when irradiated with proton beams, low and high energy X-rays and electrons. DL-alanine was impregnated with 5 nm gold-nanoparticles (3% by weight) and added as a uniform layer within a wax pellet of dimensions 10 × 5 × 5 mm. Control pellets, containing DL-Alanine were also produced, and placed within a phantom, and exposed to various types of radiations: low energy (kV ranges) X-rays were obtained from a superficial machine, high energy (MV) X-rays and electrons derived from a linear accelerator, and protons were produced by the Hyogo Ion Beam Centre in Japan. Nominal doses received ranged from 2 to 20 Gy (within clinical range). The Electron Paramagnetic Resonance (EPR) spectra of the irradiated samples were recorded on a BRUKER Elexsys 9.5 MHz. The dose enhancement caused by gold nanoparticles for 80 kV x-rays was found to be more than 60% at about 5 Gy. Smaller dose enhancements (under the same measurement conditions) were observed for megavoltage x-ray beams (up to 10%). Dose enhancement caused by charged particles indicated minimal values for 6 MeV electrons (approximately 5%) whilst less than that is obtained with protons of 150 MeV. The proton results validate the latest simulation results based on Monte Carlo calculations but the dose enhancement is significantly less than that reported in cell and animal model systems, (about 20%). We attribute this difference to the fact that alanine only measures the levels of free radicals generated by the inclusion of nanoparticles and not the redox type radicals (such as reactive oxygen species) generated from aqueous media in cells. Dose enhancement caused by 5 nm gold-nanoparticles with radiotherapy type proton beams has been found to be less than 5% as determined when using alanine/wax as both a phantom and dosimeter. This agrees well

  2. Measurement of high energy x-ray beam penumbra with Gafchromic trade mark sign EBT radiochromic film

    International Nuclear Information System (INIS)

    Cheung Tsang; Butson, Martin J.; Yu, Peter K. N.

    2006-01-01

    High energy x-ray beam penumbra are measured using Gafchromic trade mark sign EBT film. Gafchromic trade mark sign EBT, due to its limited energy dependence and high spatial resolution provide a high level of accuracy for dose assessment in penumbral regions. The spatial resolution of film detector systems is normally limited by the scanning resolution of the densitometer. Penumbral widths (80%/20%) measured at D max were found to be 2.8, 3.0, 3.2, and 3.4 mm (±0.2 mm) using 5, 10, 20, and 30 cm square field sizes, respectively, for a 6 MV linear accelerator produced x-ray beam. This is compared to 3.2 mm±0.2 mm (Kodak EDR2) and 3.6 mm±0.2 mm (Kodak X-Omat V) at 10 cmx10 cm measured using radiographic film. Using a zero volume extrapolation technique for ionization chamber measurements, the 10 cmx10 cm field penumbra at D max was measured to be 3.1 mm, a close match to Gafchromic trade mark sign EBT results. Penumbral measurements can also be made at other depths, including the surface, as the film does not suffer significantly from dosimetric variations caused by changing x-ray energy spectra. Gafchromic trade mark sign EBT film provides an adequate measure of penumbral dose for high energy x-ray beams

  3. Quantum beams

    International Nuclear Information System (INIS)

    Uesaka, Mitsuru

    2003-01-01

    Present state and future prospect are described on quantum beams for medical use. Efforts for compactness of linac for advanced cancer therapy have brought about the production of machines like Accuray's CyberKnife and TOMOTHERAPY (Tomo Therapy Inc.) where the acceleration frequency of X-band (9-11 GHz) is used. For cervical vein angiography by the X-band linac, a compact hard X-ray source is developed which is based on the (reverse) Compton scattering through laser-electron collision. More intense beam and laser are necessary at present. A compact machine generating the particle beam of 10 MeV-1 GeV (laser-plasma accelerator) for cancer therapy is also developed using the recent compression technique (chirped-pulse amplification) to generate laser of >10 TW. Tokyo University is studying for the electron beam with energy of GeV order, for the laser-based synchrotron X-ray, and for imaging by the short pulse ion beam. Development of advanced compact accelerators is globally attempted. In Japan, a virtual laboratory by National Institute of Radiological Sciences (NIRS), a working group of universities and research facilities through the Ministry of Education, Culture, Sports, Science and Technology, started in 2001 for practical manufacturing of the above-mentioned machines for cancer therapy and for angiography. Virtual Factory (Inc.), a business venture, is to be stood in future. (N.I.)

  4. X-RAY RADIATION MECHANISMS AND BEAMING EFFECT OF HOT SPOTS AND KNOTS IN ACTIVE GALACTIC NUCLEAR JETS

    International Nuclear Information System (INIS)

    Zhang Jin; Bai, J. M.; Chen Liang; Liang Enwei

    2010-01-01

    The observed radio-optical-X-ray spectral energy distributions (SEDs) of 22 hot spots and 45 knots in the jets of 35 active galactic nuclei are complied from the literature and modeled with single-zone lepton models. It is found that the observed luminosities at 5 GHz (L 5 G Hz ) and at 1 keV (L 1 k eV ) are tightly correlated, and the two kinds of sources can be roughly separated with a division of L 1 k eV = L 5 G Hz . Our SED fits show that the mechanisms of the X-rays are diverse. While the X-ray emission of a small fraction of the sources is a simple extrapolation of the synchrotron radiation for the radio-to-optical emission, an inverse Compton (IC) scattering component is necessary to model the X-rays for most of the sources. Considering the sources at rest (the Doppler factor δ = 1), the synchrotron-self-Compton (SSC) scattering would dominate the IC process. This model can interpret the X-rays of some hot spots with a magnetic field strength (B δ= 1 ssc ) being consistent with the equipartition magnetic field (B δ= 1 eq ) in 1 order of magnitude, but an unreasonably low B δ= 1 ssc is required to model the X-rays for all knots. Measuring the deviation between B δ= 1 ssc and B δ= 1 eq with ratio R B ≡ B δ= 1 eq /B δ= 1 ssc , we find that R B is greater than 1 and it is tightly anti-correlated with ratio R L ≡ L 1 k eV /L 5 G Hz for both the knots and the hot spots. We propose that the deviation may be due to the neglect of the relativistic bulk motion for these sources. Considering this effect, the IC/cosmic microwave background (CMB) component would dominate the IC process. We show that the IC/CMB model well explains the X-ray emission for most sources under the equipartition condition. Although the derived beaming factor (δ) and co-moving equipartition magnetic field (B' eq ) of some hot spots are comparable to the knots, the δ values of the hot spots tend to be smaller and their B' eq values tend to be larger than that of the knots, favoring

  5. Spectrometry of X-ray beams using Cadmium and Zinc Teluride detector

    International Nuclear Information System (INIS)

    Becker, Paulo Henriques Bastos

    1997-06-01

    Determination of X-ray spectra to be utilized for medical diagnostics is a complementary process to the development of procedures to be applied to the quality control of radiodiagnostics X-ray equipment. Until some years ago, that was only possible using Germanium or Silicon detectors. Both have an excellent resolution in this energy range, but present also some restrictions as there are high costs and the necessity of operating them at temperature of liquid Nitrogen, which is not always available at the measurement's place. Room temperature detectors like Cadmium Telluride and Mercury Iodine don't have these restrictions. They, however, have a lower resolution and incomplete collection of the charges produced by their interaction with radiation. With technological advance of crystal growth in general and new techniques like cooling the crystal with a Peltier cell and rise time discrimination circuits, today Cadmium Telluride detectors show a resolution very close to that from Germanium detectors. This work relates to the routine use of Cadmium and Zinc Telluride detectors for measuring X-ray spectra in loco of diagnostic X-ray units. It characterizes the properties of a commercially available detector and offers a model for stripping the measured pulse height distribution. It was also developed a collimator to allow the direct measurement of the beam. The model developed and the constructed set-up were applied to two X-ray tubes and the achieved spectra compared with some spectra available from the literature. (author)

  6. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... changes seen in metabolic conditions. assist in the detection and diagnosis of bone cancer . locate foreign objects ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose ...

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... current x-ray images for diagnosis and disease management. top of page How is the procedure performed? ... have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures ...

  8. Development of a Beam Trajectory Monitoring System Using e+/e- Pair Production Events

    Science.gov (United States)

    Kimura, Shota; Emoto, Yusaku; Fujihara, Kento; Ito, Hiroshi; Kawai, Hideyuki; Kobayashi, Atsushi; Mizuno, Takahiro

    2018-01-01

    In particle therapy, it is important to monitor the Bragg-peak position. It was simulated by GEANT4 Monte Carlo Simulation Code that the distribution of secondary generated gamma rays on the carbon beam therapy and the proton beam therapy. This simulation shows that gamma rays whose energy is 10 MeV or more are intensively generated at the Bragg-peak position. We are developing the system to monitor the Bragg-peak position which can measure pair production events occurred in the detector by gamma rays from irradiation points. The momentum direction of the gamma ray can be determined by measuring passing points and energy of e+ and e- generated by pair production. This system has 5 parts. The first is the conversion part. This part consists of several layers. Each layer is composed of a La-GPS ((Gd0.75La0.24Ce0.01)2Si2O7) scintillator plate and wavelength-shifting fibre (WLSF) sheets. The scintillator plate is sandwiched between sheets, where the directions of the sheets are in orthogonally x and y directions. In this part, gamma rays are converted to e+ e- pairs and the position where the conversion occured is determined. The second is the tracking part. This part consists of 2 layers of scintillating fibre tracker. Each layer has 6 scintillating fibre sheets for x, x', u, u', v, and v'. The third is the energy measurement part. It measures the energy of e+ and e- by scintillator array and Silicon Photomultipliers. The fourth is the veto counter for bremsstrahlung gamma rays from e+ and e-. The fifth is the beam monitor. By experiment, the number of photoelectrons of La-GPS with a WLSF (B-3(300)MJ, Kuraray) sheet and scintillating fibre (SCSF-78, Kuraray) when charged particle passed was measured as 9.7 and 7.6 respectively.

  9. Polycapillary x-ray optics for macromolecular crystallography

    International Nuclear Information System (INIS)

    Owens, S.M.; Gibson, W.M.; Carter, D.C.; Sisk, R.C.; Ho, J.X.

    1996-01-01

    Polycapillary x-ray optics have found potential application in many different fields, including antiscatter and magnification in mammography, radiography, x-ray fluorescence, x-ray lithography, and x-ray diffraction techniques. In x-ray diffraction, an optic is used to collect divergent x-rays from a point source and redirect them into a quasi-parallel, or slightly focused beam. Monolithic polycapillary optics have been developed recently for macromolecular crystallography and have already shown considerable gains in diffracted beam intensity over pinhole collimation. Development is being pursued through a series of simulations and prototype optics. Many improvements have been made over the stage 1 prototype reported previously, which include better control over the manufacturing process, reducing the diameter of the output beam, and addition of a slight focusing at the output of the optic to further increase x-ray flux at the sample. The authors report the characteristics and performance of the stage 1 and stage 2 optics

  10. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... body. Once it is carefully aimed at the part of the body being examined, an x-ray machine produces a small burst of radiation that passes through the body, recording an image on photographic film or a special detector. Different parts of the body absorb the x-rays in ...

  11. The superconducting x-ray lithography source program at Brookhaven

    Energy Technology Data Exchange (ETDEWEB)

    Williams, G. P.; Heese, R. N.; Vignola, G.; Murphy, J. B.; Godel, J. B.; Hsieh, H.; Galayda, J.; Seifert, A.; Knotek, M. L.

    1989-07-01

    A compact electron storage ring with superconducting dipole magnets, is being developed at the National Synchrotron Light Source at Brookhaven. The parameters of the source have been optimized for its future use as an x-ray source for lithography. This first ring is a prototype which will be used to study the operating characteristics of machines of this type with particular attention being paid to low-energy injection and long beam lifetime.

  12. Carbon ion beam is more effective to induce cell death in sphere-type A172 human glioblastoma cells compared with X-rays.

    Science.gov (United States)

    Takahashi, Momoko; Hirakawa, Hirokazu; Yajima, Hirohiko; Izumi-Nakajima, Nakako; Okayasu, Ryuichi; Fujimori, Akira

    2014-12-01

    To obtain human glioblastoma cells A172 expressing stem cell-related protein and comparison of radiosensitivity in these cells with X-rays and carbon beam. Human monolayer-type A172 glioblastoma cells were maintained in normal medium with 10% bovine serum. In order to obtain sphere-type A172 cells the medium was replaced with serum-free medium supplemented with growth factors. Both types of A172 cells were irradiated with either X-rays or carbon ion beams and their radiosensitivity was evaluated. Serum-free medium induced expression of stem cell-related proteins in A172 cells along with the neurosphere-like appearance. These sphere-type cells were found resistant to both X-rays and carbon ion beams. Phosphorylation of histone H2A family member X persisted for a longer period in the cells exposed to carbon ion beams than in those exposed to X-rays and it disappeared quicker in the sphere type than in the monolayer type. Relative radioresistance of the sphere type cells was smaller for carbon ion beams than for X-rays. We demonstrated that glioblastoma A172 cells with induced stem cell-related proteins turned resistant to irradiation. Accelerated heavy ion particles may have advantage over X-rays in overcoming the tumor resistance due to cell stemness.

  13. Pulse-periodic generation of supershort avalanche electron beams and X-ray emission

    Science.gov (United States)

    Baksht, E. Kh.; Burachenko, A. G.; Erofeev, M. V.; Tarasenko, V. F.

    2014-05-01

    Pulse-periodic generation of supershort avalanche electron beams (SAEBs) and X-ray emission in nitrogen, as well as the transition from a single-pulse mode to a pulse-periodic mode with a high repetition frequency, was studied experimentally. It is shown that, in the pulse-periodic mode, the full width at halfmaximum of the SAEB is larger and the decrease rate of the gap voltage is lower than those in the single-pulse mode. It is found that, when the front duration of the voltage pulse at a nitrogen pressure of 90 Torr decreases from 2.5 to 0.3 ns, the X-ray exposure dose in the pulse-periodic mode increases by more than one order of magnitude and the number of SAEB electrons also increases. It is shown that, in the pulse-periodic mode of a diffuse discharge, gas heating in the discharge gap results in a severalfold increase in the SAEB amplitude (the number of electrons in the beam). At a generator voltage of 25 kV, nitrogen pressure of 90 Torr, and pulse repetition frequency of 3.5 kHz, a runaway electron beam was detected behind the anode foil.

  14. X-ray generator

    Energy Technology Data Exchange (ETDEWEB)

    Zucarias, A; Shepherd, J W

    1982-09-08

    An X-ray tube has a tubular envelope with a cathode for directing an electron beam onto a focal spot area of a spaced anode target to generate X-rays. The target is mounted for axial rotation on one end of a rotor disposed in an end portion of the envelope and encircled by a stator of an alternating current induction motor. An annular shield of high permeability magnetic material extends transversely between the electron beam and the stator of the induction motor for shunting stray or fringe electromagnetic fields established by the stator away from the electron beam to avoid consequent lateral deflections of the electron and corresponding lateral movements of the focal spot area.

  15. Dental X-ray apparatus

    International Nuclear Information System (INIS)

    Weiss, M.E.

    1980-01-01

    Intra-oral X-ray apparatus which reduces the number of exposures necessary to obtain panoramic dental radiographs is described in detail. It comprises an electron gun, a tubular target carrier projecting from the gun along the beam axis and carrying at its distal end a target surrounded by a shield of X-ray opaque material. This shield extends forward and laterally of the target and has surfaces which define a wedge or cone-shaped radiation pattern delimited vertically by the root tips of the patient's teeth. A film holder is located externally of the patient's mouth. A disposable member can fit on the target carrier to depress the patient's tongue out of the radiation pattern and to further shield the roof of the mouth. The electron beam can be magnetically deflected to change the X-ray beam direction. (author)

  16. Observation of γ rays > 1015 eV from Cygnus X-3

    International Nuclear Information System (INIS)

    Lloyd-Evans, J.; Coy, R.N.; Lambert, A.; Lapikens, J.; Patel, M.; Reid, R.J.O.; Watson, A.A.

    1983-01-01

    The X-ray binary system Cygnus X-3 is a source of particular interest. As well as emitting X-rays which are modulated with a 4.8-h period, it has been observed in 30-100 MeV γ rays by the SAS II satellite and several groups have detected γ rays in the TeV range showing the same period. Most recently, using the small extensive air shower array at Kiel, workers have found that the γ-ray spectrum of Cygnus X-3 extends above 2 x 10 15 eV, with an integral flux of (7.4 +- 3.2) x 10 -14 cm -2 s -1 . This paper confirms the Kiel observations and presents evidence that the γ-ray spectrum of Cygnus X-3 steepens above 10 16 eV. (author)

  17. X-ray-ultraviolet beam splitters for the Michelson interferometer

    International Nuclear Information System (INIS)

    Delmotte, Franck; Ravet, Marie-Francoise; Bridou, Francoise; Varniere, Francoise; Zeitoun, Philippe; Hubert, Sebastien; Vanbostal, Laurent; Soullie, Gerard

    2002-01-01

    With the aim of realizing a Michelson interferometer working at 13.9 nm, we have developed a symmetrical beam splitter with multilayers deposited on the front and back sides of a silicon nitride membrane. On the basis of the experimental optical properties of the membrane, simulations have been performed to define the multilayer structure that provides the highest reflectivity-transmission product. Optimized Mo-Si multilayers have been successfully deposited on both sides of the membrane by use of the ion-beam sputtering technique, with a thickness-period reproducibility of 0.1 nm. Measurements by means of synchrotron radiation at 13.9 nm and at an angle of 45 deg. provide a reflectivity of 14.2% and a transmission of 15.2% for a 60% s-polarized light, close to the simulated values. Such a beam splitter has been used for x-ray laser Michelson interferometry at 13.9 nm. The first interferogram is discussed

  18. X-ray-ultraviolet beam splitters for the Michelson interferometer.

    Science.gov (United States)

    Delmotte, Franck; Ravet, Marie-Françoise; Bridou, Françoise; Varnière, Françoise; Zeitoun, Philippe; Hubert, Sébastien; Vanbostal, Laurent; Soullie, Gérard

    2002-10-01

    With the aim of realizing a Michelson interferometer working at 13.9 nm, we have developed a symmetrical beam splitter with multilayers deposited on the front and back sides of a silicon nitride membrane. On the basis of the experimental optical properties of the membrane, simulations have been performed to define the multilayer structure that provides the highest reflectivity-transmission product. Optimized Mo-Si multilayers have been successfully deposited on both sides of t he membrane by use of the ion-beam sputtering technique, with a thickness-period reproducibility of 0.1 nm. Measurements by means of synchrotron radiation at 13.9 nm and at an angle of 45 degrees provide a reflectivity of 14.2% and a transmission of 15.2% for a 60% s-polarized light, close to the simulated values. Such a beam splitter has been used for x-ray laser Michelson interferometry at 13.9 nm. The first interferogram is discussed.

  19. Calibration of the Gamma-RAy Polarimeter Experiment (GRAPE) at a polarized hard X-ray beam

    International Nuclear Information System (INIS)

    Bloser, P.F.; Legere, J.S.; McConnell, M.L.; Macri, J.R.; Bancroft, C.M.; Connor, T.P.; Ryan, J.M.

    2009-01-01

    The Gamma-RAy Polarimeter Experiment (GRAPE) is a concept for an astronomical hard X-ray Compton polarimeter operating in the 50-500 keV energy band. The instrument has been optimized for wide-field polarization measurements of transient outbursts from energetic astrophysical objects such as gamma-ray bursts and solar flares. The GRAPE instrument is composed of identical modules, each of which consists of an array of scintillator elements read out by a multi-anode photomultiplier tube (MAPMT). Incident photons Compton scatter in plastic scintillator elements and are subsequently absorbed in inorganic scintillator elements; a net polarization signal is revealed by a characteristic asymmetry in the azimuthal scattering angles. We have constructed a prototype GRAPE module containing a single CsI(Na) calorimeter element, at the center of the MAPMT, surrounded by 60 plastic elements. The prototype has been combined with custom readout electronics and software to create a complete 'engineering model' of the GRAPE instrument. This engineering model has been calibrated using a nearly 100% polarized hard X-ray beam at the Advanced Photon Source at Argonne National Laboratory. We find modulation factors of 0.46±0.06 and 0.48±0.03 at 69.5 and 129.5 keV, respectively, in good agreement with Monte Carlo simulations. In this paper we present details of the beam test, data analysis, and simulations, and discuss the implications of our results for the further development of the GRAPE concept.

  20. Estimating photometric redshifts for X-ray sources in the X-ATLAS field using machine-learning techniques

    Science.gov (United States)

    Mountrichas, G.; Corral, A.; Masoura, V. A.; Georgantopoulos, I.; Ruiz, A.; Georgakakis, A.; Carrera, F. J.; Fotopoulou, S.

    2017-12-01

    We present photometric redshifts for 1031 X-ray sources in the X-ATLAS field using the machine-learning technique TPZ. X-ATLAS covers 7.1 deg2 observed with XMM-Newton within the Science Demonstration Phase of the H-ATLAS field, making it one of the largest contiguous areas of the sky with both XMM-Newton and Herschel coverage. All of the sources have available SDSS photometry, while 810 additionally have mid-IR and/or near-IR photometry. A spectroscopic sample of 5157 sources primarily in the XMM/XXL field, but also from several X-ray surveys and the SDSS DR13 redshift catalogue, was used to train the algorithm. Our analysis reveals that the algorithm performs best when the sources are split, based on their optical morphology, into point-like and extended sources. Optical photometry alone is not enough to estimate accurate photometric redshifts, but the results greatly improve when at least mid-IR photometry is added in the training process. In particular, our measurements show that the estimated photometric redshifts for the X-ray sources of the training sample have a normalized absolute median deviation, nmad ≈ 0.06, and a percentage of outliers, η = 10-14%, depending upon whether the sources are extended or point like. Our final catalogue contains photometric redshifts for 933 out of the 1031 X-ray sources with a median redshift of 0.9. The table of the photometric redshifts is only available at the CDS via anonymous ftp to http://cdsarc.u-strasbg.fr (http://130.79.128.5) or via http://cdsarc.u-strasbg.fr/viz-bin/qcat?J/A+A/608/A39

  1. Combined optic system based on polycapillary X-ray optics and single-bounce monocapillary optics for focusing X-rays from a conventional laboratory X-ray source

    Energy Technology Data Exchange (ETDEWEB)

    Sun, Xuepeng; Liu, Zhiguo [The Key Laboratory of Beam Technology and Materials Modification of the Ministry of Education, Beijing Normal University, Beijing 100875 (China); College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875 (China); Beijing Radiation Center, Beijing 100875 (China); Sun, Tianxi, E-mail: stx@bnu.edu.cn [The Key Laboratory of Beam Technology and Materials Modification of the Ministry of Education, Beijing Normal University, Beijing 100875 (China); College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875 (China); Beijing Radiation Center, Beijing 100875 (China); Yi, Longtao; Sun, Weiyuan; Li, Fangzuo; Jiang, Bowen [The Key Laboratory of Beam Technology and Materials Modification of the Ministry of Education, Beijing Normal University, Beijing 100875 (China); College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875 (China); Beijing Radiation Center, Beijing 100875 (China); Ma, Yongzhong [Center for Disease Control and Prevention of Beijing, Beijing 100013 (China); Ding, Xunliang [The Key Laboratory of Beam Technology and Materials Modification of the Ministry of Education, Beijing Normal University, Beijing 100875 (China); College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875 (China); Beijing Radiation Center, Beijing 100875 (China)

    2015-12-01

    Two combined optic systems based on polycapillary X-ray optics and single-bounce monocapillary optics (SBMO) were designed for focusing the X-rays from a conventional laboratory X-ray source. One was based on a polycapillary focusing X-ray lens (PFXRL) and a single-bounce ellipsoidal capillary (SBEC), in which the output focal spot with the size of tens of micrometers of the PFXRL was used as the “virtual” X-ray source for the SBEC. The other system was based on a polycapillary parallel X-ray lens (PPXRL) and a single-bounce parabolic capillary (SBPC), in which the PPXRL transformed the divergent X-ray beam from an X-ray source into a quasi-parallel X-ray beam with the divergence of sever milliradians as the incident illumination of the SBPC. The experiment results showed that the combined optic systems based on PFXRL and SBEC with a Mo rotating anode X-ray generator with the focal spot with a diameter of 300 μm could obtain a focal spot with the total gain of 14,300 and focal spot size of 37.4 μm, and the combined optic systems based on PPXRL and SBPC with the same X-ray source mentioned above could acquire a focal spot with the total gain of 580 and focal spot size of 58.3 μm, respectively. The two combined optic systems have potential applications in micro X-ray diffraction, micro X-ray fluorescence, micro X-ray absorption near edge structure, full field X-ray microscopes and so on.

  2. Dynamical x-ray diffraction studies of interfacial strain in superlattices grown by molecular beam epitaxy

    International Nuclear Information System (INIS)

    Vandenberg, J.M.; Chu, S.N.G.; Hamm, R.A.; Panish, M.B.; Ritter, D.; Mancrander, A.T.

    1992-01-01

    This paper reports on dynamical X-ray diffraction studies that have been carried out for lattice-matched InGaAs/InP superlattices grown by modified molecular beam epitaxy (MBE) techniques. The (400) X-ray satellite pattern, which is predominantly affected by the strain modulation, was analyzed. The strain and thickness of the actual layers including the presence of strained interfacial regions were determined

  3. Development of isodose curves for A 6 mv x-ray beam

    Energy Technology Data Exchange (ETDEWEB)

    Hidaytalla, Lamya Abbas [Department of Physics, Faculty of Science, University of Khartoum, Khartoum (Sudan)

    1994-08-01

    In this thesis radiation distribution of 6 Mv x-ray beam in a water phantom is developed. The method is based on a simple empirical equation and the assumption that the x-ray source is a point source. This leads to a simple equation for the calculation and plotting of the isodose curves. the charts obtained for two fields, 6*6 and 12*12 cm{sup 2} show good agreement with the previous data used in the isotope and radiation centre, Khartoum hospital. It is suggested that further development should be carried out by writing a computer program for all the fields. (Author) 14 refs. , 9 tabs. , 34 figs. Also available from the Department of Physics, Faculty of Science, University of Khartoum, Khartoum (SD)

  4. Measurement of conversion coefficients between air Kerma and personal dose equivalent and backscatter factors for diagnostic X-ray beams

    International Nuclear Information System (INIS)

    Rosado, Paulo Henrique Goncalves

    2008-01-01

    Two sets of quantities are import in radiological protection: the protection and operational quantities. Both sets can be related to basic physical quantities such as kerma through conversion coefficients. For diagnostic x-ray beams the conversion coefficients and backscatter factors have not been determined yet, those parameters are need for calibrating dosimeters that will be used to determine the personal dose equivalent or the entrance skin dose. Conversion coefficients between air kerma and personal dose equivalent and backscatter factors were experimentally determined for the diagnostic x-ray qualities RQR and RQA recommended by the International Electrotechnical Commission (IEC). The air kerma in the phantom and the mean energy of the spectrum were measured for such purpose. Harshaw LiF-100H thermoluminescent dosemeters (TLD) were used for measurements after being calibrated against an 180 cm 3 Radcal Corporation ionization chamber traceable to a reference laboratory. A 300 mm x 300 mm x 150 mm polymethylmethacrylate (PMMA) slab phantom was used for deep-dose measurements. Tl dosemeters were placed in the central axis of the x-ray beam at 5, 10, 15, 25 and 35 mm depth in the phantom upstream the beam direction Another required parameter for determining the conversion coefficients from was the mean energy of the x-ray spectrum. The spectroscopy of x-ray beams was done with a CdTe semiconductor detector that was calibrated with 133 Ba, 241 Am and 57 Co radiation sources. Measurements of the x-ray spectra were carried out for all RQR and RQA IEC qualities. Corrections due to the detector intrinsic efficiency, total energy absorption, escape fraction of the characteristic x-rays, Compton effect and attenuation in the detector were done aiming an the accurate determination of the mean energy. Measured x-ray spectra were corrected with the stripping method by using these response functions. The typical combined standard uncertainties of conversion coefficients and

  5. Smart x-ray beam position monitor system for the Advanced Photon Source

    International Nuclear Information System (INIS)

    Shu, D.; Kuzay, T.

    1996-01-01

    In third-generation synchrotron radiation sources, such as the Advanced Photon Source (APS), the sensitivity and reliability requirements for the x-ray beam position monitors (XBPMs) are much higher than for earlier systems. Noise and contamination signals caused by radiation emitted from the bending magnet become a major problem. The regular XBPM calibration process can only provide signal correction for one set of conditions for the insertion devices (ID). During normal operation, parameters affecting the ID-emitted beam, such as the gap of the ID magnets and the beam current, are the variables. A new smart x-ray beam position monitor system (SBPM) has been conceived and designed for the APS. It has a built in self-learning structure with EEPROM memory that is large enough to open-quote open-quote remember close-quote close-quote a complete set of calibration data covering all the possible operating conditions. During the self-learning mode, the monitor system initializes a series of automatic scan motions with information for different ID setups and records them into the database array. During normal operation, the SBPM corrects the normalized output according to the ID setup information and the calibration database. So that, with this novel system, the SBPM is always calibrating itself with the changing ID set up conditions. copyright 1996 American Institute of Physics

  6. Investigation of the applicability of a special parallel-plate ionization chamber for x-ray beam dosimetry

    International Nuclear Information System (INIS)

    Perini, Ana P.; Neves, Lucio P.; Caldas, Linda V.E.

    2014-01-01

    Diagnostic x-rays are the greatest source of exposition to ionizing radiation of the population worldwide. In order to obtain accurate and lower-cost dosimeters for quality control assurance of medical x-ray facilities, a special ionization chamber was designed at the Calibration Laboratory of the IPEN, for dosimetry in diagnostic radiology beams. For the chamber characterization some tests were undertaken. Monte Carlo simulations were proposed to evaluate the distribution of the deposited energy in the sensitive volume of the ionization chamber and the collecting electrode effect on the chamber response. According to the obtained results, this special ionization chamber presents potential use for dosimetry of conventional diagnostic radiology beams. - Highlights: • An ionization chamber with a novel design was characterized for x-ray beam dosimetry. • This ionization chamber was evaluated in diagnostic radiology qualities. • The characterization tests results were within the recommended limits. • Monte Carlo simulations were employed to evaluate the design of the dosimeter. • The developed prototype is a good alternative for calibration laboratories and clinics

  7. Soft X-ray generation in gases by means of a pulsed electron beam produced in a high-voltage barier discharge

    NARCIS (Netherlands)

    Azarov, A.V.; Peters, P.J.M.; Boller, Klaus J.

    2007-01-01

    A large area pulsed electron beam is produced by a high-voltage barrier discharge. We compare the properties of the x-rays generated by stopping this beam of electrons in a thin metal foil with those generated by stopping the electrons directly in various gases. The generation of x-rays was

  8. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens

    International Nuclear Information System (INIS)

    Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K.

    2004-01-01

    A polycapillary X-ray lens is an effective optics to obtain a μm-size X-ray beam for micro-X-ray fluorescence spectrometry (μ-XRF). We developed a μ-XRF instrument using a polycapillary X-ray lens, which also enabled us to perform Grazing Exit μ-XRF (GE-μ-XRF). The evaluated diameter of the primary X-ray beam was 48 μm at the focal distance of the X-ray lens. Use of this instrument enabled two-dimensional mapping of the elemental distributions during growth of the plant 'Quinoa'. The results of the mapping revealed elemental transition during growth. In addition, a small region of thin film was analyzed by GE-μ-XRF. We expect that GE-μ-XRF will become an effective method of estimating the film thickness of a small region

  9. Three-Dimensional X-Ray Diffraction Technique for Metals Science

    DEFF Research Database (Denmark)

    Zhang, Yubin; Fan, Guohua

    2017-01-01

    The three-dimensional X-ray diffraction (3DXRD) is a new, advanced technique for materials characterization. This technique utilizes high-energy synchrotron X-rays to characterize the 3D crystallographic structure and strain/stress state of bulk materials. As the measurement is non......-destructive, the microstructural evolution as a function of time can be followed, i.e. it allows 4D (x, y, z characterizations, t). The high brilliance of synchrotron X-rays ensures that diffraction signals from volumes of micrometer scale can be quickly detected and distinguished from the background noise, i.e. its spatial...... implemented in several large synchrotron facilities, e.g. the Advanced Photon Source (APS) in USA and the Spring-8 in Japan. Another family of 3DXRD technique that utilizes white beam synchrotron X-rays has also been developed in parallel in cooperation between Oak Ridge National Laboratory and APS...

  10. X-ray tubes

    International Nuclear Information System (INIS)

    Young, R.W.

    1979-01-01

    A form of x-ray tube is described which provides satisfactory focussing of the electron beam when the beam extends for several feet from gun to target. Such a tube can be used for computerised tomographic scanning. (UK)

  11. High resolution beam profiling of X-ray free electron laser radiation by polymer imprint development.

    Science.gov (United States)

    Rösner, Benedikt; Döring, Florian; Ribič, Primož R; Gauthier, David; Principi, Emiliano; Masciovecchio, Claudio; Zangrando, Marco; Vila-Comamala, Joan; De Ninno, Giovanni; David, Christian

    2017-11-27

    High resolution metrology of beam profiles is presently a major challenge at X-ray free electron lasers. We demonstrate a characterization method based on beam imprints in poly (methyl methacrylate). By immersing the imprints formed at 47.8 eV into organic solvents, the regions exposed to the beam are removed similar to resist development in grayscale lithography. This allows for extending the sensitivity of the method by more than an order of magnitude compared to the established analysis of imprints created solely by ablation. Applying the Beer-Lambert law for absorption, the intensity distribution in a micron-sized focus can be reconstructed from one single shot with a high dynamic range, exceeding 10 3 . The procedure described here allows for beam characterization at free electron lasers revealing even faint beam tails, which are not accessible when using ablation imprint methods. We demonstrate the greatly extended dynamic range on developed imprints taken in focus of conventional Fresnel zone plates and spiral zone plates producing beams with a topological charge.

  12. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... standards used by radiology professionals. Modern x-ray systems have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that those parts of a patient's body not being imaged receive minimal radiation exposure. ...

  13. Inelastic X-ray scattering activities in Europe

    International Nuclear Information System (INIS)

    Dorner, B.

    1984-01-01

    Inelastic X-ray scattering requires an energy determination before and after the scattering process together with a technique to vary at least one energy continuously in a controlled way. Sufficiently monochromatic beams can only be produced by Bragg reflection from single crystals. Stationary X-ray monochromators are standard equipment of conventional X-ray generators to select a particular characteristic line. Quite often they are curved to focus on the sample or the detector. Devices with variable Bragg angle have been and are used as analyzers in Compton scattering which is inelastic X-ray scattering with moderate resolution. With the rapidly increasing availability of synchrotron radiation (SR) monochromators and analyzers became more and more sophisticated improving momentum (Q) resolution and only somewhat the energy resolution ΔE which stays in the order of eV. Very high energy resolution can only be obtained with Bragg angles Theta near to 90 0 . This field is the topic of the present paper

  14. Comparative effectiveness of gamma-rays and electron beams in food irradiation

    International Nuclear Information System (INIS)

    Hayashi, Toru

    1991-01-01

    Ionizing radiations which can be used for the treatment of foods are gamma-rays from Co-60 and Cs-137, accelerated electrons from a machine at an energy of 10 MeV or lower and X-rays from a machine at an energy of 5 MeV or lower. The Joint FAO/IAEA/WHO Expert Committee on the Wholesomeness of Irradiated Food held in 1980 concluded that the foods irradiated at overall average doses up to 10 kGy with the radiation listed above are wholesome for human consumption. While most of the commercial food irradiations are conducted with gamma-rays from Co-60, accelerated electrons are increasingly utilized for treating foods. An important difference between gamma-rays and accelerated electrons is the penetration capacity in materials. The penetration capacity of gamma-rays is much higher than that of accelerated electrons. Another important difference is the dose rate. The dose rates of gamma-rays from commercial Co-60 sources are 1-100 Gy/min, while those of electron beams from electron accelerators are 10 3 -10 6 Gy/s. Ideally a comparison of the effect of different types of ionizing radiation should be carried out at the same dose rate but this has been difficult due to the design of irradiators. It is very difficult to draw a definite conclusion on the difference in the effectiveness in food irradiation between gamma-rays and electron beams based on published data. This chapter deals with as many reports as possible on the comparative effectiveness of gamma-rays and electron beams and on the effect of dose rate on chemical reactions and living organisms, whether or not they demonstrate any dependency of the effect of irradiation on dose rate and type of radiation. (author)

  15. Frequency filter of seed x-ray by use of x-ray laser medium. Toward the generation of the temporally coherent x-ray laser

    International Nuclear Information System (INIS)

    Hasegawa, Noboru; Kawachi, Tetsuya; Kishimoto, Maki; Sukegawa, Kouta; Tanaka, Momoko; Ochi, Yoshihiro; Nishikino, Masaharu; Nagashima, Keisuke; Kato, Yoshiaki; Renzhong, Tai

    2009-01-01

    We evaluate the characteristics of a higher-order harmonics light as a seed X-ray amplified through a laser-produced X-ray amplifier. The narrow spectral bandwidth of the X-ray amplifier works as the frequency filter of the seed X-ray, resulting in that only the temporally coherent X-ray is amplified. Experimental investigation using the 29th-order harmonic light of the Ti:sapphire laser at a wavelength of 26.9 nm together with a neon-like manganese X-ray laser medium shows evident spectral narrowing of the seed X-ray and amplification without serious diffraction effects on the propagation of the amplified X-ray beam. This implies that the present combination is potential to realize temporally coherent X-ray lasers, with an expected duration of approximately 400 fs. (author)

  16. Application of focused-beam flat-sample method to synchrotron powder X-ray diffraction with anomalous scattering effect

    International Nuclear Information System (INIS)

    Tanaka, M; Katsuya, Y; Matsushita, Y

    2013-01-01

    The focused-beam flat-sample method (FFM), which is a method for high-resolution and rapid synchrotron X-ray powder diffraction measurements by combination of beam focusing optics, a flat shape sample and an area detector, was applied for diffraction experiments with anomalous scattering effect. The advantages of FFM for anomalous diffraction were absorption correction without approximation, rapid data collection by an area detector and good signal-to-noise ratio data by focusing optics. In the X-ray diffraction experiments of CoFe 2 O 4 and Fe 3 O 4 (By FFM) using X-rays near the Fe K absorption edge, the anomalous scattering effect between Fe/Co or Fe 2+ /Fe 3+ can be clearly detected, due to the change of diffraction intensity. The change of observed diffraction intensity as the incident X-ray energy was consistent with the calculation. The FFM is expected to be a method for anomalous powder diffraction.

  17. Excitation-resolved cone-beam x-ray luminescence tomography.

    Science.gov (United States)

    Liu, Xin; Liao, Qimei; Wang, Hongkai; Yan, Zhuangzhi

    2015-07-01

    Cone-beam x-ray luminescence computed tomography (CB-XLCT), as an emerging imaging technique, plays an important role in in vivo small animal imaging studies. However, CB-XLCT suffers from low-spatial resolution due to the ill-posed nature of reconstruction. We improve the imaging performance of CB-XLCT by using a multiband excitation-resolved imaging scheme combined with principal component analysis. To evaluate the performance of the proposed method, the physical phantom experiment is performed with a custom-made XLCT/XCT imaging system. The experimental results validate the feasibility of the method, where two adjacent nanophosphors (with an edge-to-edge distance of 2.4 mm) can be located.

  18. Kharkov X-ray Generator Based On Compton Scattering

    International Nuclear Information System (INIS)

    Shcherbakov, A.; Zelinsky, A.; Mytsykov, A.; Gladkikh, P.; Karnaukhov, I.; Lapshin, V.; Telegin, Y.; Androsov, V.; Bulyak, E.; Botman, J.I.M.; Tatchyn, R.; Lebedev, A.

    2004-01-01

    Nowadays X-ray sources based on storage rings with low beam energy and Compton scattering of intense laser beams are under development in several laboratories. An international cooperative project of an advanced X-ray source of this type at the Kharkov Institute of Physics and Technology (KIPT) is described. The status of the project is reviewed. The design lattice of the storage ring and calculated X-ray beam parameters are presented. The results of numerical simulation carried out for proposed facility show a peak spectral X-ray intensity of about 1014 can be produced

  19. X-ray facility for the ground calibration of the X-ray monitor JEM-X on board INTEGRAL

    DEFF Research Database (Denmark)

    Loffredo, G.; Pelliciari, C.; Frontera, F.

    2003-01-01

    We describe the X-ray facility developed for the calibration of the X-ray monitor JEM-X on board the INTEGRAL satellite. The apparatus allowed the scanning of the detector geometric area with a pencil beam of desired energy over the major part of the passband of the instrument. The monochromatic...

  20. Spot size characterization of focused non-Gaussian X-ray laser beams.

    Science.gov (United States)

    Chalupský, J; Krzywinski, J; Juha, L; Hájková, V; Cihelka, J; Burian, T; Vysín, L; Gaudin, J; Gleeson, A; Jurek, M; Khorsand, A R; Klinger, D; Wabnitz, H; Sobierajski, R; Störmer, M; Tiedtke, K; Toleikis, S

    2010-12-20

    We present a new technique for the characterization of non-Gaussian laser beams which cannot be described by an analytical formula. As a generalization of the beam spot area we apply and refine the definition of so called effective area (A(eff)) [1] in order to avoid using the full-width at half maximum (FWHM) parameter which is inappropriate for non-Gaussian beams. Furthermore, we demonstrate a practical utilization of our technique for a femtosecond soft X-ray free-electron laser. The ablative imprints in poly(methyl methacrylate) - PMMA and amorphous carbon (a-C) are used to characterize the spatial beam profile and to determine the effective area. Two procedures of the effective area determination are presented in this work. An F-scan method, newly developed in this paper, appears to be a good candidate for the spatial beam diagnostics applicable to lasers of various kinds.

  1. Imaging-therapy computed tomography with quasi-monochromatic X-rays.

    Science.gov (United States)

    Jost, Gregor; Golfier, Sven; Lawaczeck, Ruediger; Weinmann, Hanns-Joachim; Gerlach, Martin; Cibik, Levent; Krumrey, Michael; Fratzscher, Daniel; Rabe, Johannis; Arkadiev, Vladimir; Haschke, Michael; Langhoff, Norbert; Wedell, Reiner; Luedemann, Lutz; Wust, Peter; Pietsch, Hubertus

    2008-12-01

    Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. The X-ray optical module filters selectively the energy of the tungsten K alpha-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.

  2. SU-F-I-75: Half-Value Layer Thicknesses and Homogeneity Coefficients for Fluoroscopic X-Ray Beam Spectra Incorporating Spectral Filtration

    Energy Technology Data Exchange (ETDEWEB)

    Wunderle, K [Cleveland Clinic Foundation, Cleveland, OH (United States); Wayne State University School of Medicine, Detroit, MI (United States); Godley, A; Shen, Z; Dong, F [Cleveland Clinic Foundation, Cleveland, OH (United States); Rakowski, J [Wayne State University School of Medicine, Detroit, MI (United States)

    2016-06-15

    Purpose: The purpose of this investigation is to quantify various first half-value-layers (HVLs), second HVLs and homogeneity coefficients (HCs) for a state-of-the-art fluoroscope utilizing spectral (copper) filtration. Methods: A Radcal (Monrovia, Ca) AccuPro dosimeter with a 10×6-6 calibrated ionization chamber was used to measure air kerma for radiographic x-ray exposures made on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope operated in the service mode. The ionization chamber was centered in the x-ray beam at 72 cm from the focal spot with a source-to-image-distance of 120 cm. The collimators were introduced to limit the x-ray field to approximately 5 cm × 5 cm at the ionization chamber plane. Type-1100 aluminum filters, in 0.5 mm increments, were used to determine the HVL. Two HVL calculation methods were used, log-linear interpolation and Lambert-W interpolation as described by Mathieu [Med Phys, 38(8), 4546 (2011)]. Multiple measurements were made at 60, 80, 100, 120 kVp at spectral filtration thicknesses of 0, 0.1, 0.3, 0.6 and 0.9 mm. Results: First HVL, second HVL, and HCs are presented for the fluoroscopic x-ray beam spectra indicated above, with nearly identical results from the two interpolation methods. Accuracy of the set kVp was also determined and deviated less than 2%. First HVLs for fluoroscopic x-ray beam spectra without spectral filtration determined in our study were 7%–16% greater than previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. However, the FDA minimum HVL requirements changed since that publication, requiring larger HVLs as of 2006. Additionally, x-ray tube and generator architecture have substantially changed over the last 15 years providing different beam spectra. Conclusion: X-ray beam quality characteristics for state-of-the-art fluoroscopes with spectral filtration have not been published. This study provides reference data which will be useful for defining beam qualities encountered on

  3. Design studies for ITER x-ray diagnostics

    International Nuclear Information System (INIS)

    Hill, K.W.; Bitter, M.; von Goeler, S.; Hsuan, H.

    1995-01-01

    Concepts for adapting conventional tokamak x-ray diagnostics to the harsh radiation environment of ITER include use of grazing-incidence (GI) x-ray mirrors or man-made Bragg multilayer (ML) elements to remove the x-ray beam from the neutron beam, or use of bundles of glass-capillary x-ray ''light pipes'' embedded in radiation shields to reduce the neutron/gamma-ray fluxes onto the detectors while maintaining usable x-ray throughput. The x-ray optical element with the broadest bandwidth and highest throughput, the GI mirror, can provide adequate lateral deflection (10 cm for a deflected-path length of 8 m) at x-ray energies up to 12, 22, or 30 keV for one, two, or three deflections, respectively. This element can be used with the broad band, high intensity x-ray imaging system (XIS), the pulseheight analysis (PHA) survey spectrometer, or the high resolution Johann x-ray crystal spectrometer (XCS), which is used for ion-temperature measurement. The ML mirrors can isolate the detector from the neutron beam with a single deflection for energies up to 50 keV, but have much narrower bandwidth and lower x-ray power throughput than do the GI mirrors; they are unsuitable for use with the XIS or PHA, but they could be used with the XCS; in particular, these deflectors could be used between ITER and the biological shield to avoid direct plasma neutron streaming through the biological shield. Graded-d ML mirrors have good reflectivity from 20 to 70 keV, but still at grazing angles (<3 mrad). The efficiency at 70 keV for double reflection (10 percent), as required for adequate separation of the x-ray and neutron beams, is high enough for PHA requirements, but not for the XIS. Further optimization may be possible

  4. The IBA rhodotron TT1000: a very high power E-beam accelerator

    International Nuclear Information System (INIS)

    Abs, M.; Jongen, Y.; Poncelet, E.; Bol, J.-L.

    2004-01-01

    Due to the relatively low conversion efficiency of electrons into X-rays, the use of X-rays on an industrial basis requires high-power high-energy electron accelerators. Based on its experience acquired in the development of the Rhodotron (Nucl. Instrum. Methods B 40/41 (1989) 943; Nucl. Instrum. Methods B 79 (1993) 865), IBA has launched a few years ago a vigorous R and D program to develop such high-power electron accelerator devoted to X-ray industrial applications. This research program resulted in the TT1000 Rhodotron aimed at delivering 5 and 7 MeV electron beams with a current intensity of 100 mA. This project is now reaching its final phase with the qualification tests of the first prototype of this new machine

  5. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens

    Energy Technology Data Exchange (ETDEWEB)

    Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K. E-mail: tsuji@a-chem.eng.osaka-cu.ac.jp

    2004-08-31

    A polycapillary X-ray lens is an effective optics to obtain a {mu}m-size X-ray beam for micro-X-ray fluorescence spectrometry ({mu}-XRF). We developed a {mu}-XRF instrument using a polycapillary X-ray lens, which also enabled us to perform Grazing Exit {mu}-XRF (GE-{mu}-XRF). The evaluated diameter of the primary X-ray beam was 48 {mu}m at the focal distance of the X-ray lens. Use of this instrument enabled two-dimensional mapping of the elemental distributions during growth of the plant 'Quinoa'. The results of the mapping revealed elemental transition during growth. In addition, a small region of thin film was analyzed by GE-{mu}-XRF. We expect that GE-{mu}-XRF will become an effective method of estimating the film thickness of a small region.

  6. X-ray spectra from the Cornell Electron-Beam Ion Source (CEBIS I)

    International Nuclear Information System (INIS)

    Johnson, B.M.; Jones, K.W.; Kostroun, V.O.; Ghanbari, E.; Janson, S.W.

    1985-01-01

    Radiation emitted from the Cornell electron beam ion source (CEBIS I) has been surveyed with a Si(Li) x-ray detector. These spectra can be used to estimate backgrounds from electron bremsstrahlung and to evaluate the feasibility of atomic physics experiments using the CEBIS I source in this configuration. 1 ref., 2 figs

  7. High-intensity, subkilovolt x-ray calibration facility

    International Nuclear Information System (INIS)

    Kuckuck, R.W.; Gaines, J.L.; Ernst, R.D.

    1976-01-01

    A high-intensity subkilovolt x-ray calibration source utilizing proton-induced inner-shell atomic fluorescence of low-Z elements is described. The high photon yields and low bremsstrahlung background associated with this phenomenon are ideally suited to provide intense, nearly monoenergetic x-ray beams. The proton accelerator is a 3 mA, 300 kV Cockroft-Walton using a conventional rf hydrogen ion source. Seven remotely-selectable targets capable of heat dissipation of 5 kW/cm 2 are used to provide characteristic x-rays with energies between 100 and 1000 eV. Source strengths are of the order of 10 13 to 10 14 photons/sec. Methods of reducing spectral contamination due to hydrocarbon build-up on the target are discussed. Typical x-ray spectra (Cu-L, C-K and B-K) are shown

  8. Evaluation of secondary electron filter for removing contaminant electrons from high-energy 6 MV x-ray beam

    International Nuclear Information System (INIS)

    Kumagai, Kozo

    1988-01-01

    When using high energy X-rays, the dose increases at the skin surface and build-up region of beam contamination of secondary electrons coming out from the inner surface of the lineac head. At our radiotherapy department, many cases of external otitis from severe skin reactions, particularly resulting from whole brain irradiation of primary and metastatic brain tumors with a 6 MV X-ray lineac, have been encountered. An investigation was made of the physical aspects of a 6 MV X-ray beam using three electron filters, lead lucite, lead glass and lucite to remove secondary electrons. Transparent materials for filters should be preferable for locating the light field. The following results were obtained: 1) For removing secondary electrons, a lead lucite filter was found best. 2) The lead lucite filter proved most effective for removing secondary electrons from the area of treatment. It reduced the dose of irradiation to the skin surface and build-up region, and furthermore improved the depth dose relative to that without filters. 3) From a clinical standpoint, skin reactions such as external otitis remarkably decreased using a lead lucite filter. 4) It thus appears necessary to use a high energy X-ray with newly designed filters to reduce beam contamination of secondary electrons. (author)

  9. Analysis of a Novel Diffractive Scanning-Wire Beam Position Monitor (BPM) for Discriminative Profiling of Electron Vs. X Ray Beams

    International Nuclear Information System (INIS)

    Tatchyn, R.

    2011-01-01

    Recent numerical studies of Free Electron Lasers (FELs) operating in the Self Amplified Spontaneous Emission (SASE) regime indicate a large sensitivity of the gain to the degree of transverse overlap (and associated phase coherence) between the electron and photon beams traveling down the insertion device. Simulations of actual systems imply that accurate detection and correction for this relative loss of overlap, rather than correction for the absolute departure of the electron beam from a fixed axis, is the preferred function of an FEL amplifier's Beam Position Monitor (BPM) and corrector systems. In this note we propose a novel diffractive BPM with the capability of simultaneously detecting and resolving the absolute (and relative) transverse positions and profiles of electron and x-ray beams co-propagating through an undulator. We derive the equations governing the performance of the BPM and examine its predicted performance for the SLAC Linac Coherent Light Source (LCLS), viz., for profiling multi-GeV electron bunches co-propagating with one-to-several-hundred keV x-ray beams. Selected research and development (r and d) tasks for fabricating and testing the proposed BPM are discussed.

  10. Measurement of X-ray mass attenuation coefficient of nickel around the K-edge using synchrotron radiation based X-ray absorption study

    International Nuclear Information System (INIS)

    Roy, Bunty Rani; Rajput, Parasmani; Jha, S.N.; Nageswara Rao, A.S.

    2015-01-01

    The work presents the X-ray absorption fine structure (XAFS) technique for measuring the X-ray mass attenuation coefficient of nickel metal foil in the X-ray energy range of 8271.2–8849.4 eV using scanning XAFS beam line (BL-09) at Indus-2 synchrotron radiation source facility, Raja Ramanna Centre for Advanced Technology (RRCAT) at Indore, India. The result represents the X-ray mass attenuation coefficient data for 0.02 mm thick Ni metal foil in the XAFS region of Ni K-edge. However, the results are compared to theoretical values using X-COM. There is a maximum deviation which is found exactly near the K-edge jump and decreases as we move away from the absorption edge. Oscillatory structure appears just above the observed absorption edge i.e., 8348.7 eV and is confined to around 250 eV above the edge. - Highlights: • Mass attenuation coefficient measurements of nickel using synchrotron radiation. • The measurements were taken exactly near the Ni K-edge at an energy step of 1 eV. • A maximum deviation is found near the K-edge

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... have very controlled x-ray beams and dose control methods to minimize stray (scatter) radiation. This ensures that those parts of a patient's body not being imaged receive minimal radiation exposure. top of page What are the limitations of Bone X-ray (Radiography)? ...

  12. Control console for the X-ray room

    International Nuclear Information System (INIS)

    Garcia H, J.M.; Aguilar B, M.A.; Torres B, M.A.

    1998-01-01

    It is presented the design and construction of Control console for the X-ray room of Metrology Center for ionizing radiations at National Institute of Nuclear Research (ININ). This system controls the positioning of 6 different filters for an X-ray beam. Also it controls a shutter which blockades the beam during periods established by user, these periods can be fixed from hours until tenth of second. The shutter opening periods, as well as the X-ray beam filter are establish and monitoring from a Personal computer outside of room. (Author)

  13. Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy

    International Nuclear Information System (INIS)

    Ding, George X; Duggan, Dennis M; Coffey, Charles W

    2007-01-01

    The purpose of this investigation is to characterize the beams produced by a kilovoltage (kV) imager integrated into a linear accelerator (Varian on-board imager integrated into the Trilogy accelerator) for acquiring high resolution volumetric cone-beam computed tomography (CBCT) images of the patient on the treatment table. The x-ray tube is capable of generating photon spectra with kVp values between 40 and 125 kV. The Monte Carlo simulations were used to study the characteristics of kV beams and the properties of imaged target scatters. The Monte Carlo results were benchmarked against measurements, and excellent agreements were obtained. We also studied the effect of including the electron impact ionization (EII), and the simulation showed that the characteristic radiation is increased significantly in the energy spectra when EII is included. Although only slight beam hardening is observed in the spectra of all photons after passing through the phantom target, there is a significant difference in the spectra and angular distributions between scattered and primary photons. The results also show that the photon fluence distributions are significantly altered by adding bow tie filters. The results indicate that a combination of large cone-beam field size and large imaged target significantly increases scatter-to-primary ratios for photons that reach the detector panel. For phantoms 10 cm, 20 cm and 30 cm thick of water placed at the isocentre, the scatter-to-primary ratios are 0.94, 3.0 and 7.6 respectively for an open 125 kVp CBCT beam. The Monte Carlo simulations show that the increase of the scatter is proportional to the increase of the imaged volume, and this also applies to scatter-to-primary ratios. This study shows both the magnitude and the characteristics of scattered x-rays. The knowledge obtained from this investigation may be useful in the future design of the image detector to improve the image quality

  14. Multilayer X-ray mirrors for formation of sub-nanometer wavelength range beams

    International Nuclear Information System (INIS)

    Akhsakhalyan, A.A.; Akhsakhalyan, A.D.; Klyuenkov, E.B.; Murav'ev, V.A.; Salashchenko, N.N.; Kharitonov, A.I.

    2005-01-01

    Paper reviews the efforts undertaken in the RF Academy of Sciences IPM within recent 5 years to design multilayer mirror systems to produce X-ray wavelength subnanometer range beams. Paper describes a process to fabricate the mentioned systems covering the procedures to obtain supersmooth surfaces of the specified shape, to deposit gradient multilayer structures on the mentioned surfaces and describes the rules to calculate the optimal parameters of mirrors. Paper presents characteristics of mirror system two types: a mirror in the shape of a parabolic cylinder to collimate radiation in the DRON-4, DRON-6 production-type X-ray diffractometers and in the shape of a quadraelliptic reflector - a new wide-aperture four-corner focusing system [ru

  15. Electron beam production and characterization for the PLEIADES Thomson X-ray source

    International Nuclear Information System (INIS)

    Brown, W.J.; Hartemann, F.V.; Tremaine, A.M.; Springer, P.T.; Le Sage, G.P.; Barty, C.P.J.; Crane, J.K.; Cross, R.R.; Fittinghoff, D.N.; Slaughter, D.R.; Rosenzweig, J.B.; Anderson, S.; Gibson, D.J.

    2002-01-01

    We report on the performance of an S-band RF photocathode electron gun and accelerator for operation with the PLEIADES Thomson x-ray source at LLNL. Simulations of beam production, transport, and focus are presented. It is shown that a 1 ps, 500 pC electron bunch with a normalized emittance of less than 5 πmm-mrad can be delivered to the interaction point. Initial electron measurements are presented. Calculations of expected x-ray flux are also performed, demonstrating an expected peak spectral brightness of 1020 photons/s/mm2/mrad2/0.1% bandwidth. Effects of RF phase jitter are also presented, and planned phase measurements and control methods are discussed

  16. X-ray beam-shaping via deformable mirrors: Analytical computation of the required mirror profile

    International Nuclear Information System (INIS)

    Spiga, Daniele; Raimondi, Lorenzo; Svetina, Cristian; Zangrando, Marco

    2013-01-01

    X-ray mirrors with high focusing performances are in use in both mirror modules for X-ray telescopes and in synchrotron and FEL (Free Electron Laser) beamlines. A degradation of the focus sharpness arises in general from geometrical deformations and surface roughness, the former usually described by geometrical optics and the latter by physical optics. In general, technological developments are aimed at a very tight focusing, which requires the mirror profile to comply with the nominal shape as much as possible and to keep the roughness at a negligible level. However, a deliberate deformation of the mirror can be made to endow the focus with a desired size and distribution, via piezo actuators as done at the EIS-TIMEX beamline of FERMI@Elettra. The resulting profile can be characterized with a Long Trace Profilometer and correlated with the expected optical quality via a wavefront propagation code. However, if the roughness contribution can be neglected, the computation can be performed via a ray-tracing routine, and, under opportune assumptions, the focal spot profile (the Point Spread Function, PSF) can even be predicted analytically. The advantage of this approach is that the analytical relation can be reversed; i.e., from the desired PSF the required mirror profile can be computed easily, thereby avoiding the use of complex and time-consuming numerical codes. The method can also be suited in the case of spatially inhomogeneous beam intensities, as commonly experienced at synchrotrons and FELs. In this work we expose the analytical method and the application to the beam shaping problem

  17. 120 YEARS SINCE THE DISCOVERY OF X-RAYS.

    Science.gov (United States)

    Babic, Rade R; Stankovic Babic, Gordana; Babic, Strahinja R; Babic, Nevena R

    2016-09-01

    This paper is intended to celebrate the 120th anniversary of the discovery of X-rays. X-rays (Roentgen-rays) were discovered on the 8th ofNovember, 1895 by the German physicist Wilhelm Conrad Roentgen. Fifty days after the discovery of X-ray, on December 28, 1895. Wilhelm Conrad Roentgen published a paper about the discovery of X-rays - "On a new kind of rays" (Wilhelm Conrad Roentgen: Ober eine neue Art von Strahlen. In: Sitzungsberichte der Wurzburger Physik.-Medic.- Gesellschaft. 1895.). Therefore, the date of 28th ofDecember, 1895 was taken as the date of X-rays discovery. This paper describes the work of Wilhelm Conrad Roentgen, Nikola Tesla, Mihajlo Pupin and Maria Sklodowska-Curie about the nature of X-rays . The fantastic four - Wilhelm Conrad Roentgen, NikolaTesla, Mihajlo ldvorski Pupin and Maria Sklodowska-Curie set the foundation of radiology with their discovery and study of X-rays. Five years after the discovery of X-rays, in 1900, Dr Avram Vinaver had the first X-ray machine installed in abac, in Serbia at the time when many developed countries did not have an X-ray machine and thus set the foundation of radiology in Serbia.

  18. Present status of research and development on X-ray microprobe

    International Nuclear Information System (INIS)

    Koike, Masaki; Suzuki, I.H.

    1991-01-01

    X-ray beam micro-analysis has advanced rapidly in these years in conjunction with the development of powerful X-ray sources. Among a variety of methods being attempted, the method using a collimated narrow beam has been important because of high brightness, and of usability in both regions of soft and hard X-rays. In the soft X-ray region, the focused beam is formed by a fresnel zone plate or a Schwaltzschild mirror assembly, and can be used for scanning transmission microscope or scanning photoelectron microscope. In the hard X-ray region, the beam is formed by grazing incidence mirrors, and can be used for X-ray fluorescence micro-analysis for obtaining elemental mapping. In this report, the recent progress on the soft X-ray scanning microscopy and the X-ray microprobe has been surveyed, together with the improvement on the related optical elements. (author) 84 refs

  19. Time resolved, 2-D hard X-ray imaging of relativistic electron-beam target interactions on ETA-II

    International Nuclear Information System (INIS)

    Crist, C.E.; Sampayan, S.; Westenskow, G.; Caporaso, G.; Houck, T.; Weir, J.; Trimble, D.; Krogh, M.

    1998-01-01

    Advanced radiographic applications require a constant source size less than 1 mm. To study the time history of a relativistic electron beam as it interacts with a bremsstrahlung converter, one of the diagnostics they use is a multi-frame time-resolved hard x-ray camera. They are performing experiments on the ETA-II accelerator at Lawrence Livermore National Laboratory to investigate details of the electron beam/converter interactions. The camera they are using contains 6 time-resolved images, each image is a 5 ns frame. By starting each successive frame 10 ns after the previous frame, they create a 6-frame movie from the hard x-rays produced from the interaction of the 50-ns electron beam pulse

  20. Confocal total reflection X-ray fluorescence technology based on an elliptical monocapillary and a parallel polycapillary X-ray optics.

    Science.gov (United States)

    Zhu, Yu; Wang, Yabing; Sun, Tianxi; Sun, Xuepeng; Zhang, Xiaoyun; Liu, Zhiguo; Li, Yufei; Zhang, Fengshou

    2018-07-01

    A total reflection X-ray fluorescence (TXRF) spectrometer based on an elliptical monocapillary X-ray lens (MXRL) and a parallel polycapillary X-ray lens (PPXRL) was designed. This TXRF instrument has micro focal spot, low divergence and high intensity of incident X-ray beam. The diameter of the focal spot of MXRL was 16.5 µm, and the divergence of the incident X-ray beam was 3.4 mrad. We applied this TXRF instrument to the micro analysis of a single-layer film containing Ni deposited on a Si substrate by metal vapor vacuum arc ion source. Copyright © 2018 Elsevier Ltd. All rights reserved.

  1. Synchrotron X-ray micro-beam studies of ancient Egyptian make-up

    International Nuclear Information System (INIS)

    Martinetto, P.; Anne, M.; Dooryhee, E.; Drakopoulos, M.; Dubus, M.; Salomon, J.; Simionovici, A.; Walter, Ph.

    2001-01-01

    Vases full of make-up are most often present in the burial furniture of Egyptian tombs dated from the pharaonic period. The powdered cosmetics made of isolated grains are analysed to identify their trace element signature. From this signature we identify the provenance of the mineral ingredients in the make-up and we observe different impurities in products, which have been demonstrated as synthetic substances by previous works. Focused X-ray micro-beam (2x5 μm 2 ) is successively tuned at 11 keV, below the L III absorption edge of Pb, and 31.8 keV for global characterisation of the metal impurities. The fluorescence signal integrated over each single grain is detected against the X-ray micro-diffraction pattern collected in transmission with a bi-dimensional detector. Furthermore, for galena grains rich in Zn, the XANES signal at the K-absorption edge of Zn shows its immediate nearest-neighbour environment

  2. Synchrotron X-ray micro-beam studies of ancient Egyptian make-up

    Energy Technology Data Exchange (ETDEWEB)

    Martinetto, P; Anne, M; Dooryhee, E; Drakopoulos, M; Dubus, M; Salomon, J; Simionovici, A; Walter, Ph

    2001-07-01

    Vases full of make-up are most often present in the burial furniture of Egyptian tombs dated from the pharaonic period. The powdered cosmetics made of isolated grains are analysed to identify their trace element signature. From this signature we identify the provenance of the mineral ingredients in the make-up and we observe different impurities in products, which have been demonstrated as synthetic substances by previous works. Focused X-ray micro-beam (2x5 {mu}m{sup 2}) is successively tuned at 11 keV, below the L{sub III} absorption edge of Pb, and 31.8 keV for global characterisation of the metal impurities. The fluorescence signal integrated over each single grain is detected against the X-ray micro-diffraction pattern collected in transmission with a bi-dimensional detector. Furthermore, for galena grains rich in Zn, the XANES signal at the K-absorption edge of Zn shows its immediate nearest-neighbour environment.

  3. Installation And Test Of Electron Beam Generation System To Produce Far-Infrared Radiation And X-Ray Pulses

    International Nuclear Information System (INIS)

    Wichaisirimongkol, Pathom; Jinamoon, Witoon; Khangrang, Nopadon; Kusoljariyakul, Keerati; Rhodes, Michael W.; Rimjaem, Sakhorn; Saisut, Jatuporn; Chitrlada, Thongbai; Vilaithong, Thiraphat; Wiedemann, Helmut

    2005-10-01

    SURIYA project at the Fast Neutron Research Facility, Chiang Mai University, aims to establish a facility to generate femtosecond electron beams. This electron beam can be used to generate high intensity far-infrared radiation and ultra-short X-ray pulses. The main components of the system are a 3 MeV RF electron gun with a thermionic cathode, an a-magnet as a bunch compressor, and post acceleration 15-20 MeV by a linear accelerator (linac). Between the main components, there are focusing quadrupole magnets and steering magnets to maintain the electron beam within a high vacuum tube. At the end of the beam transport line, a dipole magnet has been installed to function as a beam dump and an energy spectrometer. After the installation and testing of individual major components were completed, we have been investigating the generation of the electron beam, intense far- infrared radiation and ultra short X-ray pulses

  4. Time-resolved x-ray line emission studies of thermal transport in multiple beam uv-irradiated targets

    International Nuclear Information System (INIS)

    Jaanimagi, P.A.; Henke, B.L.; Delettrez, J.; Richardson, M.C.

    1984-01-01

    Thermal transport in spherical targets irradiated with multiple, nanosecond duration laser beams, has been a topic of much discussion recently. Different inferences on the level of thermal flux inhibition have been drawn from plasma velocity and x-ray spectroscopic diagnostics. We present new measurements of thermal transport on spherical targets made through time-resolved x-ray spectroscopic measurements of the progress of the ablation surface through thin layers of material on the surface of the target. These measurements, made with 6 and 12 uv (351 nm) nanosecond beams from OMEGA, will be compared to previous thermal transport measurements. Transparencies of the conference presentation are given

  5. Output Beam Polarisation of X-ray Lasers with Transient Inversion

    Science.gov (United States)

    Janulewicz, K. A.; Kim, C. M.; Matouš, B.; Stiel, H.; Nishikino, M.; Hasegawa, N.; Kawachi, T.

    It is commonly accepted that X-ray lasers, as the devices based on amplified spontaneous emission (ASE), did not show any specific polarization in the output beam. The theoretical analysis within the uniform (single-mode) approximation suggested that the output radiation should show some defined polarization feature, but randomly changing from shot-to-shot. This hypothesis has been verified by experiment using traditional double-pulse scheme of transient inversion. Membrane beam-splitter was used as a polarization selector. It was found that the output radiation has a significant component of p-polarisation in each shot. To explain the effect and place it in the line with available, but scarce data, propagation and kinetic effects in the non-uniform plasma have been analysed.

  6. Parabolic refractive X-ray lenses: a breakthrough in X-ray optics

    CERN Document Server

    Lengeler, B; Benner, B; Guenzler, T F; Kuhlmann, M; Tümmler, J; Simionovici, A S; Drakopoulos, M; Snigirev, A; Snigireva, I

    2001-01-01

    Refractive X-ray lenses, considered for a long time as unfeasible, have been realized with a rotational parabolic profile at our institute: The main features of the new lenses are: they focus in two directions and are free of spherical aberration. By varying the number of individual lenses in the stack the focal length can be chosen in a typical range from 0.5 to 2 m for photon energies between about 6 and 60 keV. The aperture of the lens is about 1 mm matching the angular divergence of undulator beams at 3d generation synchrotron radiation sources. They cope without problems with the heat load from the white beam of an undulator. Finally, they are easy to align and to operate. Refractive X-ray lenses can be used with hard X-rays in the same way as glass lenses can be used for visible light, if it is take into account that the numerical aperture is small (of the order 10 sup - sup 4). Being high-quality optical elements, the refractive X-ray lenses can be used for generating a focal spot in the mu m range wit...

  7. Soft X-ray generation via inverse compton scattering between high quality electron beam and high power laser

    International Nuclear Information System (INIS)

    Masakazu Washio; Kazuyuki Sakaue; Yoshimasa Hama; Yoshio Kamiya; Tomoko Gowa; Akihiko Masuda; Aki Murata; Ryo Moriyama; Shigeru Kashiwagi; Junji Urakawa

    2007-01-01

    High quality beam generation project based on High-Tech Research Center Project, which has been approved by Ministry of Education, Culture, Sports, Science and Technology in 1999, has been conducted by advance research institute for science and engineering, Waseda University. In the project, laser photo-cathode RF-gun has been selected for the high quality electron beam source. RF cavities with low dark current, which were made by diamond turning technique, have been successfully manufactured. The low emittance electron beam was realized by choosing the modified laser injection technique. The obtained normalized emmitance was about 3 m.mrad at 100 pC of electron charge. The soft x-ray beam generation with the energy of 370 eV, which is in the energy region of so-called water window, by inverse Compton scattering has been performed by the collision between IR laser and the low emmitance electron beams. (Author)

  8. Geometrical E-beam proximity correction for raster scan systems

    Science.gov (United States)

    Belic, Nikola; Eisenmann, Hans; Hartmann, Hans; Waas, Thomas

    1999-04-01

    High pattern fidelity is a basic requirement for the generation of masks containing sub micro structures and for direct writing. Increasing needs mainly emerging from OPC at mask level and x-ray lithography require a correction of the e-beam proximity effect. The most part of e-beam writers are raster scan system. This paper describes a new method for geometrical pattern correction in order to provide a correction solution for e-beam system that are not able to apply variable doses.

  9. Feasibility of X-ray analysis of multi-layer thin films at a single beam voltage

    International Nuclear Information System (INIS)

    Statham, P J

    2010-01-01

    Multi-layer analysis using electron beam excitation and X-ray spectrometry is a powerful tool for characterising layers down to 1 nm thickness and with typically 1 μm lateral resolution but does not always work. Most published applications have used WDS with many measurements at different beam voltages and considerable experience has been needed to choose lines and voltages particularly for complex multi-layer problems. A new objective mathematical approach is described which demonstrates whether X-ray analysis can obtain reliable results for an arbitrary multi-layer problem. A new algorithm embodied in 'ThinFilmID' software produces a single plot that shows feasibility of achieving results with a single EDS spectrum and suggests the optimal beam voltage. Synthesis of EDS spectra allows the precision in results to be estimated and acquisition conditions modified before wasting valuable instrument time. Thus, practicality of multi-layer thin film analysis at a single beam voltage can now be established without the extensive experimentation that was previously required by a microanalysis expert. Examples are shown where the algorithm discovers viable single-voltage conditions for applications that experts previously thought could only be addressed using measurements at more than one beam voltage.

  10. Sampling conditions of the three dimensional (3D) fan beam X ray transform

    International Nuclear Information System (INIS)

    Desbat, L.; Roux, S.; Grangeat, P.; Koenig, A.

    2003-01-01

    We give the sampling conditions of the 3 D fan beam X ray transform (3.D.F.B.X.R.T.). The motivation of this work lie in the fact that helical tomography with a single detector line is simply a sampling of this transform under the helical constraint. we give a precise description of the geometry of the essential support of Fourier transform of the 3.D.F.B.X.R.T. and we show how to derive efficient sampling. (N.C.)

  11. Compton backscattered collmated X-ray source

    Science.gov (United States)

    Ruth, Ronald D.; Huang, Zhirong

    2000-01-01

    A high-intensity, inexpensive and collimated x-ray source for applications such as x-ray lithography is disclosed. An intense pulse from a high power laser, stored in a high-finesse resonator, repetitively collides nearly head-on with and Compton backscatters off a bunched electron beam, having relatively low energy and circulating in a compact storage ring. Both the laser and the electron beams are tightly focused and matched at the interaction region inside the optical resonator. The laser-electron interaction not only gives rise to x-rays at the desired wavelength, but also cools and stabilizes the electrons against intrabeam scattering and Coulomb repulsion with each other in the storage ring. This cooling provides a compact, intense bunch of electrons suitable for many applications. In particular, a sufficient amount of x-rays can be generated by this device to make it an excellent and flexible Compton backscattered x-ray (CBX) source for high throughput x-ray lithography and many other applications.

  12. Compton backscattered collimated x-ray source

    Science.gov (United States)

    Ruth, R.D.; Huang, Z.

    1998-10-20

    A high-intensity, inexpensive and collimated x-ray source is disclosed for applications such as x-ray lithography is disclosed. An intense pulse from a high power laser, stored in a high-finesse resonator, repetitively collides nearly head-on with and Compton backscatters off a bunched electron beam, having relatively low energy and circulating in a compact storage ring. Both the laser and the electron beams are tightly focused and matched at the interaction region inside the optical resonator. The laser-electron interaction not only gives rise to x-rays at the desired wavelength, but also cools and stabilizes the electrons against intrabeam scattering and Coulomb repulsion with each other in the storage ring. This cooling provides a compact, intense bunch of electrons suitable for many applications. In particular, a sufficient amount of x-rays can be generated by this device to make it an excellent and flexible Compton backscattered x-ray (CBX) source for high throughput x-ray lithography and many other applications. 4 figs.

  13. National Synchrotron Light Source users manual: Guide to the VUV and x-ray beam lines

    International Nuclear Information System (INIS)

    Gmuer, N.F.; White-DePace, S.M.

    1987-08-01

    The success of the National Synchrotron Light Source in the years to come will be based, in large part, on the size of the users community and the diversity of the scientific disciplines represented by these users. In order to promote this philosophy, this National Synchrotron Light Source (NSLS) Users Manual: Guide to the VUV and X-Ray Beam Lines, has been published. This manual serves a number of purposes. In an effort to attract new research, it will present to the scientific community-at-large the current and projected architecture and capabilities of the various VUV and x-ray beam lines and storage rings. We anticipate that this publication will be updated periodically in order to keep pace with the constant changes at the NSLS

  14. High-resolution accelerator alignment using x-ray optics

    Directory of Open Access Journals (Sweden)

    Bingxin Yang

    2006-03-01

    Full Text Available We propose a novel alignment technique utilizing the x-ray beam of an undulator in conjunction with pinholes and position-sensitive detectors for positioning components of the accelerator, undulator, and beam line in an x-ray free-electron laser. Two retractable pinholes at each end of the undulator define a stable and reproducible x-ray beam axis (XBA. Targets are precisely positioned on the XBA using a pinhole camera technique. Position-sensitive detectors responding to both x-ray and electron beams enable direct transfer of the position setting from the XBA to the electron beam. This system has the potential to deliver superior alignment accuracy (1–3   μm for target pinholes in the transverse directions over a long distance (200 m or longer. It can be used to define the beam axis of the electron-beam–based alignment, enabling high reproducibility of the latter. This x-ray–based concept should complement the electron-beam–based alignment and the existing survey methods to raise the alignment accuracy of long accelerators to an unprecedented level. Further improvement of the transverse accuracy using x-ray zone plates will be discussed. We also propose a concurrent measurement scheme during accelerator operation to allow real-time feedback for transverse position correction.

  15. Self-compensating x-ray or γ-ray thickness gauge

    International Nuclear Information System (INIS)

    Allport, J.J.

    1977-01-01

    A gauge is described for determining the mass per unit area, or alternatively the thickness of sheet material by measuring the attenuation, as well as backscatter, of an x-ray beam or the like, while continuously taking into account deviations and changes in localized material composition, insofar as these have an effect on the transmission coefficient of the beam. Electrical signals representing these deviations are combined with calibration data for given material nominal properties, i.e., nominal composition. The resultant and output signal represents the mass per unit area or thickness

  16. Ultrashort X-ray pulse science

    Energy Technology Data Exchange (ETDEWEB)

    Chin, Alan Hap [Univ. of California, Berkeley, CA (US). Dept. of Physics; Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)

    1998-05-01

    A variety of phenomena involves atomic motion on the femtosecond time-scale. These phenomena have been studied using ultrashort optical pulses, which indirectly probe atomic positions through changes in optical properties. Because x-rays can more directly probe atomic positions, ultrashort x-ray pulses are better suited for the study of ultrafast structural dynamics. One approach towards generating ultrashort x-ray pulses is by 90° Thomson scattering between terawatt laser pulses and relativistic electrons. Using this technique, the author generated ~ 300 fs, 30 keV (0.4 Å) x-ray pulses. These x-ray pulses are absolutely synchronized with ultrashort laser pulses, allowing femtosecond optical pump/x-ray probe experiments to be performed. Using the right-angle Thomson scattering x-ray source, the author performed time-resolved x-ray diffraction studies of laser-perturbated InSb. These experiments revealed a delayed onset of lattice expansion. This delay is due to the energy relaxation from a dense electron-hole plasma to the lattice. The dense electron-hole plasma first undergoes Auger recombination, which reduces the carrier concentration while maintaining energy content. Longitudinal-optic (LO) phonon emission then couples energy to the lattice. LO phonon decay into acoustic phonons, and acoustic phonon propagation then causes the growth of a thermally expanded layer. Source characterization is instrumental in utilizing ultrashort x-ray pulses in time-resolved x-ray spectroscopies. By measurement of the electron beam diameter at the generation point, the pulse duration of the Thomson scattered x-rays is determined. Analysis of the Thomson scattered x-ray beam properties also provides a novel means of electron bunch characterization. Although the pulse duration is inferred for the Thomson scattering x-ray source, direct measurement is required for other x-ray pulse sources. A method based on the laser-assisted photoelectric effect (LAPE) has been demonstrated as a

  17. Ultrashort X-ray pulse science

    International Nuclear Information System (INIS)

    Chin, A.H.; Lawrence Berkeley National Lab., CA

    1998-01-01

    A variety of phenomena involves atomic motion on the femtosecond time-scale. These phenomena have been studied using ultrashort optical pulses, which indirectly probe atomic positions through changes in optical properties. Because x-rays can more directly probe atomic positions, ultrashort x-ray pulses are better suited for the study of ultrafast structural dynamics. One approach towards generating ultrashort x-ray pulses is by 90 o Thomson scattering between terawatt laser pulses and relativistic electrons. Using this technique, the author generated ∼ 300 fs, 30 keV (0.4 (angstrom)) x-ray pulses. These x-ray pulses are absolutely synchronized with ultrashort laser pulses, allowing femtosecond optical pump/x-ray probe experiments to be performed. Using the right-angle Thomson scattering x-ray source, the author performed time-resolved x-ray diffraction studies of laser-perturbated InSb. These experiments revealed a delayed onset of lattice expansion. This delay is due to the energy relaxation from a dense electron-hole plasma to the lattice. The dense electron-hole plasma first undergoes Auger recombination, which reduces the carrier concentration while maintaining energy content. Longitudinal-optic (LO) phonon emission then couples energy to the lattice. LO phonon decay into acoustic phonons, and acoustic phonon propagation then causes the growth of a thermally expanded layer. Source characterization is instrumental in utilizing ultrashort x-ray pulses in time-resolved x-ray spectroscopies. By measurement of the electron beam diameter at the generation point, the pulse duration of the Thomson scattered x-rays is determined. Analysis of the Thomson scattered x-ray beam properties also provides a novel means of electron bunch characterization. Although the pulse duration is inferred for the Thomson scattering x-ray source, direct measurement is required for other x-ray pulse sources. A method based on the laser-assisted photoelectric effect (LAPE) has been

  18. Development of a Reference System for the determination of the personal dose equivalent and the constancy of X- Ray beams

    International Nuclear Information System (INIS)

    Vivolo, Vitor

    2006-01-01

    A reference system for the determination of the personal dose equivalent, Hp (10), and a quality control program of X-ray equipment used In radioprotection require the periodic verification of the X-ray beams constancy. In this work, two parallel-plate ionization chambers were developed with inner electrodes of different materials, and inserted into PMMA slab phantoms. One ionization chamber was developed with inner carbon electrodes and the other with inner aluminium electrodes. The two ionization chambers can be used as a Tandem system. The different energy response of the two ionization chambers allowed the development of the Tandem system that is very useful for the checking of the constancy of beam qualities. Standard intermediary energy X-ray beams (from 48 keV to 118 keV), radioprotection level, were established through the development of a dosimetric methodology and the analysis of their physical parameters. The ionization chambers were studied in relation to their operational characteristics, and they were calibrated in X-ray beams (radioprotection, diagnostic radiology, mammography and radiotherapy levels) in accordance to international recommendations. They presented good performance. The determination procedure of personal dose equivalent, Hp (10), was established. (author)

  19. Characteristics and quality test of X-ray with CZT detector

    International Nuclear Information System (INIS)

    Rianto, Sugeng

    2000-01-01

    The study examines the use of direct measurements of x-ray spectra for testing the quality of x-ray beam by using a peltier-cooled CZT detector under different conditions. The typical calibration of the spectrometry system shows that the energy resolution of the system is 1,2 keV at 122 keV of 57 Co. The utilization CZT based spectrometer for assessing the quality of x-ray machine on its spectra show that the CZT could accurately measure the spectra at various kVp, m As and filtration, except at the kVp greater than 140 and m As higher than 2 without added filtration. A comparison of CZT with the field instruments showed that there is a reasonable agreement between the Keithley and CZT at lower energies regardless of filtration however at high energies there is a large difference.In contrast the discrepancy between the CZT and Nero increased at lower photon energies particularly for high filtration

  20. E-line: A new crystal collimator beam line for source size measurements at CHESS

    International Nuclear Information System (INIS)

    White, Jeffrey A.; Revesz, Peter; Finkelstein, Ken

    2007-01-01

    A new X-ray beam line has been constructed at cornell high energy synchrotron source (CHESS) to measure the vertical and horizontal source size of the positron particle beam. The cornell laboratory of elementary particle physics (LEPP) operates the storage ring (CESR) for X-ray generation for the CHESS user community by circulating electrons and their antimatter counterpart positrons in counter-rotating beams. As the laboratory reduces the emittances of particle beams to increase X-ray brilliance, there has been an increasing need for diagnostic tools to measure and monitor source size. A beam line front end that accesses the positron synchrotron light has been fitted with an experimental chamber and apparatus of compact design capable of horizontal and vertical source size measurement using the 'crystal collimator' technique, and an additional setup for vertical beam position monitoring using a luminescence-based X-ray video beam position monitoring system. The crystal collimators each consist of two Si(2 2 0) crystals in a dispersive (+,+) arrangement that diffract X-rays to a fluorescent material coated on a view port observed with a CCD camera. Measurements of the positron vertical beam size using the crystal collimation method at E-line are compared with measurements of visible synchrotron light at a remotely located dedicated port on the storage ring

  1. Studies of a Linac Driver for a High Repetition Rate X-Ray FEL

    International Nuclear Information System (INIS)

    Venturini, M.; Corlett, J.; Doolittle, L.; Filippetto, D.; Papadopoulos, C.; Penn, G.; Prosnitz, D.; Qiang, J.; Reinsch, M.; Ryne, R.; Sannibale, F.; Staples, J.; Wells, R.; Wurtele, J.; Zolotorev, M.; Zholents, A.

    2011-01-01

    We report on on-going studies of a superconducting CW linac driver intended to support a high repetition rate FEL operating in the soft x-rays spectrum. We present a pointdesign for a 1.8 GeV machine tuned for 300 pC bunches and delivering low-emittance, low-energy spread beams as needed for the SASE and seeded beamlines.

  2. A one-dimensional ion beam figuring system for x-ray mirror fabrication

    International Nuclear Information System (INIS)

    Idir, Mourad; Huang, Lei; Bouet, Nathalie; Kaznatcheev, Konstantine; Vescovi, Matthew; Lauer, Ken; Conley, Ray; Rennie, Kent; Kahn, Jim; Nethery, Richard; Zhou, Lin

    2015-01-01

    We report on the development of a one-dimensional Ion Beam Figuring (IBF) system for x-ray mirror polishing. Ion beam figuring provides a highly deterministic method for the final precision figuring of optical components with advantages over conventional methods. The system is based on a state of the art sputtering deposition system outfitted with a gridded radio frequency inductive coupled plasma ion beam source equipped with ion optics and dedicated slit developed specifically for this application. The production of an IBF system able to produce an elongated removal function rather than circular is presented in this paper, where we describe in detail the technical aspect and present the first obtained results

  3. A one-dimensional ion beam figuring system for x-ray mirror fabrication

    Energy Technology Data Exchange (ETDEWEB)

    Idir, Mourad, E-mail: midir@bnl.gov; Huang, Lei; Bouet, Nathalie; Kaznatcheev, Konstantine; Vescovi, Matthew; Lauer, Ken [NSLS-II, Brookhaven National Laboratory, P.O. Box 5000, Upton, New York 11973 (United States); Conley, Ray [Advanced Photon Source, Argonne National Laboratory, Argonne, Illinois 60439 (United States); Rennie, Kent; Kahn, Jim; Nethery, Richard [Kaufman & Robinson, Inc., 1330 Blue Spruce Drive, Fort Collins, Colorado 80524 (United States); Zhou, Lin [College of Mechatronics and Automation, National University of Defense Technology, 109 Deya Road, Changsha, Hunan 410073 (China); Hu’nan Key Laboratory of Ultra-precision Machining Technology, Changsha, Hunan 410073 (China)

    2015-10-15

    We report on the development of a one-dimensional Ion Beam Figuring (IBF) system for x-ray mirror polishing. Ion beam figuring provides a highly deterministic method for the final precision figuring of optical components with advantages over conventional methods. The system is based on a state of the art sputtering deposition system outfitted with a gridded radio frequency inductive coupled plasma ion beam source equipped with ion optics and dedicated slit developed specifically for this application. The production of an IBF system able to produce an elongated removal function rather than circular is presented in this paper, where we describe in detail the technical aspect and present the first obtained results.

  4. Improving Patient Safety with X-Ray and Anesthesia Machine Ventilator Synchronization: A Medical Device Interoperability Case Study

    Science.gov (United States)

    Arney, David; Goldman, Julian M.; Whitehead, Susan F.; Lee, Insup

    When a x-ray image is needed during surgery, clinicians may stop the anesthesia machine ventilator while the exposure is made. If the ventilator is not restarted promptly, the patient may experience severe complications. This paper explores the interconnection of a ventilator and simulated x-ray into a prototype plug-and-play medical device system. This work assists ongoing interoperability framework development standards efforts to develop functional and non-functional requirements and illustrates the potential patient safety benefits of interoperable medical device systems by implementing a solution to a clinical use case requiring interoperability.

  5. Multilayer on-chip stacked Fresnel zone plates: Hard x-ray fabrication and soft x-ray simulations

    Energy Technology Data Exchange (ETDEWEB)

    Li, Kenan; Wojcik, Michael J.; Ocola, Leonidas E.; Divan, Ralu; Jacobsen, Chris

    2015-11-01

    Fresnel zone plates are widely used as x-ray nanofocusing optics. To achieve high spatial resolution combined with good focusing efficiency, high aspect ratio nanolithography is required, and one way to achieve that is through multiple e-beam lithography writing steps to achieve on-chip stacking. A two-step writing process producing 50 nm finest zone width at a zone thickness of 1.14 µm for possible hard x-ray applications is shown here. The authors also consider in simulations the case of soft x-ray focusing where the zone thickness might exceed the depth of focus. In this case, the authors compare on-chip stacking with, and without, adjustment of zone positions and show that the offset zones lead to improved focusing efficiency. The simulations were carried out using a multislice propagation method employing Hankel transforms.

  6. Multilayer optics for x-ray analysis: design - fabrication - application

    International Nuclear Information System (INIS)

    Dietsch, R.; Holz, Th.; Bruegemann, L.

    2002-01-01

    substrate dimensions. Magnetron sputtering and e-beam evaporation are well established deposition techniques to fabricate X-ray optical multilayers. For specific layer material combinations and tailored thickness profiles, Pulsed Laser Deposition (PLD) has become an interesting alternative to these predominant technologies. As a result of the accuracy achieved with PLD, gradient multilayers of different material combinations can be deposited both on flat and on pre-curved substrates. For several years, Ni/C- Goebel-Mirrors are well established in X-ray diffraction using Cu Kα-radiation. Intensities of more then 10 9 cps together with a low beam divergence Δφ 1 ) : l(Cu Kβ) > 10 6 are realized with the Twin Goebel-Mirror arrangement (TGM). The TGM arrangement is also available for Mo Kα-radiation. A combination of a graded multilayer mirror for Mo Kα-radiation and a Germanium channel cut monochromator is capable of providing monochromatic Mo Kα 1 - radiation. There are further applications of multilayer optics in the high energy range for synchrotron, medical and astrophysical investigations. In surface analyses like AES and SIMS, these nanometer-multilayers can also be used to study depth resolution effects in depth profiling analysis. Copyright (2002) Australian X-ray Analytical Association Inc

  7. Imaging-therapy computed tomography with quasi-monochromatic X-rays

    International Nuclear Information System (INIS)

    Jost, Gregor; Golfier, Sven; Lawaczeck, Ruediger; Weinmann, Hanns-Joachim; Gerlach, Martin; Cibik, Levent; Krumrey, Michael; Fratzscher, Daniel; Rabe, Johannis; Arkadiev, Vladimir; Haschke, Michael; Langhoff, Norbert; Wedell, Reiner

    2008-01-01

    Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. Results: The X-ray optical module filters selectively the energy of the tungsten Kα-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT

  8. Imaging-therapy computed tomography with quasi-monochromatic X-rays

    Energy Technology Data Exchange (ETDEWEB)

    Jost, Gregor [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: gregor.jost@bayerhealthcare.com; Golfier, Sven [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: sven.golfier@bayerhealthcare.com; Lawaczeck, Ruediger [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: ruediger.lawaczeck@bayerhealthcare.com; Weinmann, Hanns-Joachim [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: hanns-joachim.weinmann@bayerhealthcare.com; Gerlach, Martin [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: martin.gerlach@ptb.de; Cibik, Levent [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: levent.cibik@ptb.de; Krumrey, Michael [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: michael.krumrey@ptb.de; Fratzscher, Daniel [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Fratzscher@ifg-adlershof.de; Rabe, Johannis [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Rabe@ifg-adlershof.de; Arkadiev, Vladimir [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Arkadiev@ifg-adlershof.de; Haschke, Michael [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Haschke@ifg-adlershof.de; Langhoff, Norbert [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Langhoff@ifg-adlershof.de; Wedell, Reiner [Institut fuer angewandte Photonik e.V., Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: wedell-iap@ifg-adlershof.de (and others)

    2008-12-15

    Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. Results: The X-ray optical module filters selectively the energy of the tungsten K{alpha}-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.

  9. X-ray pulse wavefront metrology using speckle tracking

    International Nuclear Information System (INIS)

    Berujon, Sebastien; Ziegler, Eric; Cloetens, Peter

    2015-01-01

    The theoretical description and experimental implementation of a speckle-tracking-based instrument which permits the characterisation of X-ray pulse wavefronts. An instrument allowing the quantitative analysis of X-ray pulsed wavefronts is presented and its processing method explained. The system relies on the X-ray speckle tracking principle to accurately measure the phase gradient of the X-ray beam from which beam optical aberrations can be deduced. The key component of this instrument, a semi-transparent scintillator emitting visible light while transmitting X-rays, allows simultaneous recording of two speckle images at two different propagation distances from the X-ray source. The speckle tracking procedure for a reference-less metrology mode is described with a detailed account on the advanced processing schemes used. A method to characterize and compensate for the imaging detector distortion, whose principle is also based on speckle, is included. The presented instrument is expected to find interest at synchrotrons and at the new X-ray free-electron laser sources under development worldwide where successful exploitation of beams relies on the availability of an accurate wavefront metrology

  10. Micronuclei in human peripheral blood lymphocytes exposed to mixed beams of X-rays and alpha particles

    Czech Academy of Sciences Publication Activity Database

    Staaf, E.; Brehwens, K.; Haghdoost, S.; Nievaart, S.; Pachnerová Brabcová, Kateřina; Czub, J.; Braziewicz, J.; Wojcik, A.

    2012-01-01

    Roč. 51, č. 3 (2012), s. 283-293 ISSN 0301-634X Institutional research plan: CEZ:AV0Z10480505 Keywords : Micronuclei * LET * Combined exposure * Mixed beams * Alpha particles * X-rays Subject RIV: BO - Biophysics Impact factor: 1.754, year: 2012

  11. High-energy X-ray diffraction studies of short- and intermediate-range structure in oxide glasses

    International Nuclear Information System (INIS)

    Suzuya, Kentaro

    2002-01-01

    The feature of high-energy X-ray diffraction method is explained. The oxide glasses studies by using BL04B2, high-energy X-ray diffraction beam line of SPring-8, and the random system materials by high-energy monochromatic X-ray diffraction are introduced. An advantage of third generation synchrotron radiation is summarized. On SPring-8, the high-energy X-ray diffraction experiments of random system are carried out by BL04B2 and BL14B1 beam line. BL04B2 can select Si (111)(E=37.8 keV, λ=0.033 nm) and Si(220)(E=61.7 keV, λ=0.020 nm) as Si monochromator. The intermediate-range structure of (MgO) x (P 2 O 5 ) 1-x glass ,MgP 2 O 6 glass, B 2 O 3 glass, SiO 2 and GeO 2 are explained in detail. The future and application of high-energy X-ray diffraction are stated. (S.Y.)

  12. Toward a fourth-generation X-ray source

    International Nuclear Information System (INIS)

    Monction, D. E.

    1999-01-01

    The field of synchrotron radiation research has grown rapidly over the last 25 years due to both the push of the accelerator and magnet technology that produces the x-ray beams and the pull of the extraordinary scientific research that is possible with them. Three successive generations of synchrotrons radiation facilities have resulted in beam brilliances 11 to 12 orders of magnitude greater than the standard laboratory x-ray tube. However, greater advances can be easily imagined given the fact that x-ray beams from present-day facilities do not exhibit the coherence or time structure so familiar with the optical laser. Theoretical work over the last ten years or so has pointed to the possibility of generating hard x-ray beams with laser-like characteristics. The concept is based on self-amplified spontaneous emission (SASE) in flee-electron lasers. A major facility of this type based upon a superconducting linac could produce a cost-effective facility that spans wave-lengths from the ultraviolet to the hard x-ray regime, simultaneously servicing large numbers experimenters from a wide range of disciplines. As with each past generation of synchrotrons facilities, immense new scientific opportunities would result from fourth-generation sources.

  13. X-ray spectroscopy of hydrogen-like ions in an electron beam ion trap

    Energy Technology Data Exchange (ETDEWEB)

    Tarbutt, M.R.; Crosby, D.; Silver, J.D. [Univ. of Oxford, Clarendon Lab. (United Kingdom); Myers, E.G. [Dept. of Physics, Florida State Univ., Tallahassee, FL (United States); Nakamura, N.; Ohtani, S. [ICORP, JST, Chofu, Tokyo (Japan)

    2001-07-01

    The X-ray emission from highly charged hydrogen-like ions in an electron beam ion trap is free from the problems of satellite contamination and Doppler shifts inherent in fast-beam sources. This is a favourable situation for the measurement of ground-state Lamb shifts in these ions. We present recent progress toward this goal, and discuss a method whereby wavelength comparison between transitions in hydrogenlike ions of different nuclear charge Z, enable the measurement of QED effects without requiring an absolute calibration.

  14. Synchrotron x-ray fluorescence and extended x-ray absorption fine structure analysis

    International Nuclear Information System (INIS)

    Chen, J.R.; Gordon, B.M.; Hanson, A.L.; Jones, K.W.; Kraner, H.W.; Chao, E.C.T.; Minkin, J.A.

    1984-01-01

    The advent of dedicated synchrotron radiation sources has led to a significant increase in activity in many areas of science dealing with the interaction of x-rays with matter. Synchrotron radiation provides intense, linearly polarized, naturally collimated, continuously tunable photon beams, which are used to determine not only the elemental composition of a complex, polyatomic, dilute material but also the chemical form of the elements with improved accuracy. Examples of the application of synchrotron radiation include experiments in synchrotron x-ray fluorescence (SXRF) analysis and extended x-ray absorption fine structure (EXAFS) analysis. New synchrotron radiation x-ray microprobes for elemental analysis in the parts per billion range are under construction at several laboratories. 76 references, 24 figures

  15. Absorbed dose determination in water in medium energy x-ray beam

    International Nuclear Information System (INIS)

    Nisevic, G.; Spasic-Jokic, V.

    1998-01-01

    Absorbed dose determination in water phantom in medium energy X-ray beam, according to IAEA recommendations is given. This method is applied on Radiotherapy department of Military Academy Hospital in Belgrade. Reference points of measurements are on depth of 5 cm and 2 cm as it recommended in ref. Experimental results are shown in aim to introduce new dosimetric concept based on air kerma calibration factor recommended for application in our radiotherapy centers (author)

  16. Modeling x-ray data for the Saturn z-pinch machine

    International Nuclear Information System (INIS)

    Matuska, W.; Peterson, D.; Deeney, C.; Derzon, M.

    1997-01-01

    A wealth of XRD and time dependent x-ray imaging data exist for the Saturn z-pinch machine, where the load is either a tungsten wire array or a tungsten wire array which implodes onto a SiO 2 foam. Also, these pinches have been modeled with a 2-D RMHD Eulerian computer code. In this paper the authors start with the 2-D Eulerian results to calculate time and spatially dependent spectra using both LTE and NLTE models. Then using response functions, these spectra are converted to XRD currents and camera images, which are quantitatively compared with the data. Through these comparisons, areas of good and lesser quality agreement are determined, and areas are identified where the 2-D Eulerian code should be improved

  17. Establishment of standard x-ray qualities to be used in diagnostic level at SSDLs

    International Nuclear Information System (INIS)

    A-Aziz, R.O.

    2007-08-01

    The objective of the present work is to modify x-ray machine (used by the secondary standard dosimetry laboratory in Sudan) to produce x-ray qualities suitable for the calibration at diagnostic level. It based on experimental evaluations. The importance of this study appears in development of protocols to calibrate the instruments used in diagnostic radiology in by good response of the instruments and low cost. In particular, the half value layer (HVL) values for the following qualities 40, 60, 80, 100, 120 and 150 were determined using various attenuation layers. Ionization chamber was used to determine the free air kerma rate at a distance of 100 cm from the x-ray generator. The obtained HVL values were compared with standard values of diagnostic levels. It was observed that the HVL of the present x-ray machine in most qualities are smaller than the standard ones. An approved method (described in the standard IEC 61267) was applied to determine the amount of additional filtration required to meet the standard values. The results show that the available cupper layers were not suitable for this purpose as most of the x-ray was absorbed in the material. Aluminum layers, on the other hand , show good performances to reduce the beams to the desired levels. The amount of additional filtration (from aluminum layers) needed in order to establish standard x-ray qualities to be used in diagnostic level were determined.(Author)

  18. X-ray Microprobe for Fluorescence and Diffraction Analysis

    International Nuclear Information System (INIS)

    Ice, G.E.

    2005-01-01

    X-ray diffraction (see unit 1.1) and x-ray excited fluorescence analysis are powerful techniques for the nondestructive measurement of crystal structure and chemical composition. X-ray fluorescence analysis is inherently nondestructive with orders of magnitude lower power deposited for the same detectable limit as with fluorescence excited by charged particle probes (Sparks, 1980). X-ray diffraction analysis is sensitive to crystal structure with orders-of-magnitude greater sensitivity to crystallographic strain than electron probes (Rebonato, et al. 1989). When a small-area x-ray microbeam is used as the probe, chemical composition (Z>14), crystal structure, crystalline texture, and crystalline strain distributions can be determined. These distributions can be studied both at the surface of the sample and deep within the sample (Fig. 1). Current state-of-the-art can achieve an ∼1 mm-D x-ray microprobe and an ∼0.1 mm-D x-ray microprobe has been demonstrated (Bilderback, et al., 1994). Despite their great chemical and crystallographic sensitivities, x-ray microprobe techniques have until recently been restricted by inefficient x-ray focusing optics and weak x-ray sources; x-ray microbeam analysis was largely superseded by electron techniques in the 50's. However, interest in x-ray microprobe techniques has now been revived (Howells, et al., 1983; Ice and Sparks, 1984; Chevallier, et al., 1997; Riekel 1992; Thompson, el al., 1992; and Making and Using... 1997) by the development of efficient x-ray focusing optics and ultra-high intensity synchrotron x-ray sources (Buras and Tazzari, 1984; Shenoy, et al., 1988). These advances have increased the achievable microbeam flux by more than 11 orders of magnitude (Fig. 2) (Ice, 1997); the flux in a tunable 1 mm-D beam on a 'so called' 3rd-generation synchrotron source such as the APS can exceed the flux in a fixed-energy mm2 beam on a conventional source. These advances make x-ray microfluorescence and x-ray

  19. Relativistic self-focusing of ultra-high intensity X-ray laser beams in warm quantum plasma with upward density profile

    International Nuclear Information System (INIS)

    Habibi, M.; Ghamari, F.

    2014-01-01

    The results of a numerical study of high-intensity X-ray laser beam interaction with warm quantum plasma (WQP) are presented. By means of an upward ramp density profile combined with quantum factors specially the Fermi velocity, we have demonstrated significant relativistic self-focusing (RSF) of a Gaussian electromagnetic beam in the WQP where the Fermi temperature term in the dielectric function is important. For this purpose, we have considered the quantum hydrodynamics model that modifies refractive index of inhomogeneous WQPs with the inclusion of quantum correction through the quantum statistical and diffraction effects in the relativistic regime. Also, to better illustration of the physical difference between warm and cold quantum plasmas and their effect on the RSF, we have derived the envelope equation governing the spot size of X-ray laser beam in Q-plasmas. In addition to the upward ramp density profile, we have found that the quantum effects would be caused much higher oscillation and better focusing of X-ray laser beam in the WQP compared to that of cold quantum case. Our computational results reveal the importance of the use of electrons density profile and Fermi speed in enhancing self-focusing of laser beam

  20. Flash X-Ray (FXR) Accelerator Optimization Electronic Time-Resolved Measurement of X-Ray Source Size

    International Nuclear Information System (INIS)

    Jacob, J; Ong, M; Wargo, P

    2005-01-01

    Lawrence Livermore National Laboratory (LLNL) is currently investigating various approaches to minimize the x-ray source size on the Flash X-Ray (FXR) linear induction accelerator in order to improve x-ray flux and increase resolution for hydrodynamic radiography experiments. In order to effectively gauge improvements to final x-ray source size, a fast, robust, and accurate system for measuring the spot size is required. Timely feedback on x-ray source size allows new and improved accelerator tunes to be deployed and optimized within the limited run-time constraints of a production facility with a busy experimental schedule; in addition, time-resolved measurement capability allows the investigation of not only the time-averaged source size, but also the evolution of the source size, centroid position, and x-ray dose throughout the 70 ns beam pulse. Combined with time-resolved measurements of electron beam parameters such as emittance, energy, and current, key limiting factors can be identified, modeled, and optimized for the best possible spot size. Roll-bar techniques are a widely used method for x-ray source size measurement, and have been the method of choice at FXR for many years. A thick bar of tungsten or other dense metal with a sharp edge is inserted into the path of the x-ray beam so as to heavily attenuate the lower half of the beam, resulting in a half-light, half-dark image as seen downstream of the roll-bar; by measuring the width of the transition from light to dark across the edge of the roll-bar, the source size can be deduced. For many years, film has been the imaging medium of choice for roll-bar measurements thanks to its high resolution, linear response, and excellent contrast ratio. Film measurements, however, are fairly cumbersome and require considerable setup and analysis time; moreover, with the continuing trend towards all-electronic measurement systems, film is becoming increasingly difficult and expensive to procure. Here, we shall

  1. Development of a 3-D X-ray system

    Science.gov (United States)

    Evans, James Paul Owain

    The interpretation of standard two-dimensional x-ray images by humans is often very difficult. This is due to the lack of visual cues to depth in an image which has been produced by transmitted radiation. The solution put forward in this research is to introduce binocular parallax, a powerful physiological depth cue, into the resultant shadowgraph x-ray image. This has been achieved by developing a binocular stereoscopic x-ray imaging technique, which can be used for both visual inspection by human observers and also for the extraction of three-dimensional co-ordinate information. The technique is implemented in the design and development of two experimental x-ray systems and also the development of measurement algorithms. The first experimental machine is based on standard linear x-ray detector arrays and was designed as an optimum configuration for visual inspection by human observers. However, it was felt that a combination of the 3-D visual inspection capability together with a measurement facility would enhance the usefulness of the technique. Therefore, both a theoretical and an empirical analysis of the co-ordinate measurement capability of the machine has been carried out. The measurement is based on close-range photogrammetric techniques. The accuracy of the measurement has been found to be of the order of 4mm in x, 3mm in y and 6mm in z. A second experimental machine was developed and based on the same technique as that used for the first machine. However, a major departure has been the introduction of a dual energy linear x-ray detector array which will allow, in general, the discrimination between organic and inorganic substances. The second design is a compromise between ease of visual inspection for human observers and optimum three-dimensional co-ordinate measurement capability. The system is part of an on going research programme into the possibility of introducing psychological depth cues into the resultant x-ray images. The research presented in

  2. Providing x-rays

    International Nuclear Information System (INIS)

    Mallozzi, P.J.; Epstein, H.M.

    1985-01-01

    This invention provides an apparatus for providing x-rays to an object that may be in an ordinary environment such as air at approximately atmospheric pressure. The apparatus comprises: means (typically a laser beam) for directing energy onto a target to produce x-rays of a selected spectrum and intensity at the target; a fluid-tight enclosure around the target; means for maintaining the pressure in the first enclosure substantially below atmospheric pressure; a fluid-tight second enclosure adjoining the first enclosure, the common wall portion having an opening large enough to permit x-rays to pass through but small enough to allow the pressure reducing means to evacuate gas from the first enclosure at least as fast as it enters through the opening; the second enclosure filled with a gas that is highly transparent to x-rays; the wall of the second enclosure to which the x-rays travel having a portion that is highly transparent to x-rays (usually a beryllium or plastic foil), so that the object to which the x-rays are to be provided may be located outside the second enclosure and adjacent thereto and thus receive the x-rays substantially unimpeded by air or other intervening matter. The apparatus is particularly suited to obtaining EXAFS (extended x-ray fine structure spectroscopy) data on a material

  3. Machine recognition of navel orange worm damage in x-ray images of pistachio nuts

    Science.gov (United States)

    Keagy, Pamela M.; Parvin, Bahram; Schatzki, Thomas F.

    1995-01-01

    Insect infestation increases the probability of aflatoxin contamination in pistachio nuts. A non- destructive test is currently not available to determine the insect content of pistachio nuts. This paper uses film X-ray images of various types of pistachio nuts to assess the possibility of machine recognition of insect infested nuts. Histogram parameters of four derived images are used in discriminant functions to select insect infested nuts from specific processing streams.

  4. New x-ray optical system for fluorescence beamline at Hasylab

    International Nuclear Information System (INIS)

    Falkenberg, G.; Tschentscher, T.

    2000-01-01

    Beamline L at HASYLAB/DESY is actually dedicated to micro x-ray fluorescence (μ-XRF) experiments using the white beam from a bending magnet of the storage ring DORIS III. To extend the applicability of beamline L to other x-ray fluorescence techniques, such as synchrotron radiation total reflection x-ray fluorescence (SR-TXRF) and micro x-ray absorption near edge structures in fluorescence mode (μ-XANES), new x-ray optics have been designed and are under installation at the moment. The suitability of beamline L for SR-TXRF experiments has been shown previously in a number of studies using temporary setups for beam monochromatization and collimation. The new optical system comprises a slit system, a pair of x-ray mirrors for focussing, collimation and high energy cut-off (12 keV and 30 keV), a double multilayer monochromator for broad bandpass applications (TXRF) and a double perfect-crystal monochromator for spectroscopy (XANES, speciation). The multilayer monochromator will utilize a pair of NiC with a spacing of 4.0 nm for the energy range 2-10 keV and a second pair of WB 4 C with a spacing of 3.0 nm for the range 4-30 keV. To extend the energy range for broad bandpass applications to higher photon energies SiGe gradient crystals are foreseen (ΔE/E ∼ 10 -3 ). For the perfect-crystal monochromator we have chosen a pair of Ge 111 crystals for the energy range 2-10 keV and Si 111 crystals for 7-90 keV. To enable the use of low photon energies down to 2 keV the monochromator vessel is sealed to the ring vacuum by a 25 μm thick carbon window. The mirrors and monochromators deflect the beam vertically and can be moved out of the beam independently. Fixed exit geometry permits the illumination of the same sample spot with different wavelength and energy bands. All optical elements accept the full vertical beam opening in order to enable both vertical and horizontal geometries for sample and detector. (author)

  5. X-ray generator based on Compton scattering

    NARCIS (Netherlands)

    Androsov, V.P.; Agafonov, A.V.; Botman, J.I.M.; Bulyak, E.V.; Drebot, I.; Gladkikh, P.I.; Grevtsev, V.; Ivashchenko, V.; Karnaukhov, I.M.; Lapshin, V.I.

    2005-01-01

    Nowadays, the sources of the X-rays based on a storage ring with low beam energy and Compton scattering of intense laser beam are under development in several laboratories. In the paper the state-of-art in development and construction of cooperative project of a Kharkov advanced X-ray source NESTOR

  6. Scattered X-ray beam nondestructive testing

    International Nuclear Information System (INIS)

    Harding, G.; Kosanetzky, J.

    1988-01-01

    X-ray scatter interactions generally dominate the linear attenuation coefficient at the photon energies typical of medical and industrial radiography. Specific advantages of X-ray scatter imaging, including a flexible choice of measurement geometry, direct 3D-imaging capability (tomography) and improved information for material characterization, are illustrated with results from Compton and coherent scatter devices. Applications of a Compton backscatter scanner (ComScan) in the aerospace industry and coherent scatter imaging in security screening are briefly considered [pt

  7. Development of a Hard X-ray Beam Position Monitor for Insertion Device Beams at the APS

    Science.gov (United States)

    Decker, Glenn; Rosenbaum, Gerd; Singh, Om

    2006-11-01

    Long-term pointing stability requirements at the Advanced Photon Source (APS) are very stringent, at the level of 500 nanoradians peak-to-peak or better over a one-week time frame. Conventional rf beam position monitors (BPMs) close to the insertion device source points are incapable of assuring this level of stability, owing to mechanical, thermal, and electronic stability limitations. Insertion device gap-dependent systematic errors associated with the present ultraviolet photon beam position monitors similarly limit their ability to control long-term pointing stability. We report on the development of a new BPM design sensitive only to hard x-rays. Early experimental results will be presented.

  8. Spatial resolution in depth-controlled surface sensitive x-ray techniques

    International Nuclear Information System (INIS)

    Yun, W.B.; Viccaro, P.J.

    1992-01-01

    The spatial resolution along the surface normal and the total depth probed are two important parameters in depth-controlled surface sensitive X-ray techniques employing grazing incidence geometry. The two parameters are analyzed in terms of optical properties (refractive indices) of the media involved and parameters of the incident X-ray beam: beam divergence, X-ray energy, and spectral bandwidth. We derive analytical expressions of the required beam divergence and spectral bandwidth of the incident beam as a function of the two parameters. Sample calculations are made for X-ray energies between 0.1 and 100 keV and for solid Be, Cu, and Au, representing material matrices consisting of low, medium, and high atomic number elements. A brief discussion on obtaining the required beam divergence and spectral bandwidth from present X-ray sources and optics is given

  9. SU-D-209-02: Percent Depth Dose Curves for Fluoroscopic X-Ray Beam Qualities Incorporating Copper Filtration

    Energy Technology Data Exchange (ETDEWEB)

    Wunderle, K [Cleveland Clinic Foundation, Cleveland, OH (United States); Wayne State University School of Medicine, Detroit, MI (United States); Godley, A; Shen, Z; Dong, F [Cleveland Clinic Foundation, Cleveland, OH (United States); Rakowski, J [Wayne State University School of Medicine, Detroit, MI (United States)

    2016-06-15

    Purpose: The purpose of this investigation was to quantify percent depth dose (PDD) curves for fluoroscopic x-ray beam qualities incorporating added copper filtration. Methods: A PTW (Freiburg, Germany) MP3 water tank was used with a Standard Imaging (Middleton, WI) Exradin Model 11 Spokas Chamber to measure PDD curves for 60, 80, 100 and 120 kVp x-ray beams with copper filtration ranging from 0.0–0.9 mm at 22cm and 42cm fields of view from 0 to 150 mm of water. A free-in-air monitor chamber was used to normalize the water tank data to fluctuations in output from the fluoroscope. The measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope. The fluoroscope was inverted from the typical orientation providing an x-ray beam originating from above the water tank. The water tank was positioned so that the water level was located at 60cm from the focal spot; which also represents the focal spot to interventional reference plane distance for that fluoroscope. Results: PDDs for 60, 80, 100, and 120 kVp with 0 mm of copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)] for those beam qualities given differences in fluoroscopes, geometric orientation, type of ionization chamber, and the water tank used for data collection. PDDs for 60, 80, 100, and 120 kVp with copper filtration were obtained and are presented, which have not been previously investigated and published. Conclusion: The equipment and processes used to acquire the reported data were sound and compared well with previously published data for PDDs without copper filtration. PDD data for the fluoroscopic x-ray beams incorporating copper filtration can be used as reference data for estimating organ or soft tissue dose at depth involving similar beam qualities or for comparison with mathematical models.

  10. Radiation processing with high-energy X-rays

    International Nuclear Information System (INIS)

    Cleland, Marshall R.; Stichelbaut, Frederic

    2009-01-01

    The physical, chemical or biological characteristics of selected commercial products and materials can be improved by radiation processing. The ionizing energy can be provided by accelerated electrons with energies between 75 keV and 10 MeV, gamma rays from cobalt-60 with average energies of 1.25 MeV or X-rays with maximum energies up to 7.5 MeV. Electron beams are preferred for thin products, which are processed at high speeds. Gamma rays are used for products that are too thick for treatment with electron beams. High-energy X-rays can also be used for these purposes because their penetration in solid materials is similar to or even slightly greater than that of gamma rays. Previously, the use of X-rays had been inhibited by their slower processing rates and higher costs when compared with gamma rays. Since then, the price of cobalt-60 sources has been increased and the radiation intensity from high-energy, high-power X-ray generators has also increased. For facilities requiring at least 2 MCi of cobalt-60, the capital and operating costs of X-ray facilities with equivalent processing rates can be less than that of gamma-ray irradiators. Several high-energy electron beam facilities have been equipped with removable X-ray targets so that irradiation processes can be done with either type of ionizing energy. A new facility is now being built which will be used exclusively in the X-ray mode to sterilize medical products. Operation of this facility will show that high-energy, high-power X-ray generators are practical alternatives to large gamma-ray sources. (author)

  11. Monitoring moisture movements in building materials using x-ray attenuation

    DEFF Research Database (Denmark)

    Pease, Bradley Justin; Scheffler, Gregor A.; Janssen, Hans

    2012-01-01

    analysis with a composite model consisting of a dry porous material and a thickness of water equivalent to the moisture content of the material. The current formulation of this composite model relies on certain assumptions, including a monochromatic x-ray photon beam source (i.e., x-ray photons of a single....... Implications of this inconsistency are introduced and discussed. This paper presents both an overview of fundamental descriptions of the x-ray attenuation measurement technique and results from a parametric experimental study of various porous construction materials, including calcium silicate board, aerated...

  12. Synchrotron X-ray imaging applied to solar photovoltaic silicon

    International Nuclear Information System (INIS)

    Lafford, T A; Villanova, J; Plassat, N; Dubois, S; Camel, D

    2013-01-01

    Photovoltaic (PV) cell performance is dictated by the material of the cell, its quality and purity, the type, quantity, size and distribution of defects, as well as surface treatments, deposited layers and contacts. A synchrotron offers unique opportunities for a variety of complementary X-ray techniques, given the brilliance, spectrum, energy tunability and potential for (sub-) micron-sized beams. Material properties are revealed within in the bulk and at surfaces and interfaces. X-ray Diffraction Imaging (X-ray Topography), Rocking Curve Imaging and Section Topography reveal defects such as dislocations, inclusions, misorientations and strain in the bulk and at surfaces. Simultaneous measurement of micro-X-Ray Fluorescence (μ-XRF) and micro-X-ray Beam Induced Current (μ-XBIC) gives direct correlation between impurities and PV performance. Together with techniques such as microscopy and Light Beam Induced Current (LBIC) measurements, the correlation between structural properties and photovoltaic performance can be deduced, as well as the relative influence of parameters such as defect type, size, spatial distribution and density (e.g [1]). Measurements may be applied at different stages of solar cell processing in order to follow the evolution of the material and its properties through the manufacturing process. Various grades of silicon are under study, including electronic and metallurgical grades in mono-crystalline, multi-crystalline and mono-like forms. This paper aims to introduce synchrotron imaging to non-specialists, giving example results on selected solar photovoltaic silicon samples.

  13. Degradation of organophosphorus compounds by X-ray irradiation

    International Nuclear Information System (INIS)

    Trebse, P.; Arcon, I.

    2002-01-01

    Complete text of publication follows. Our research presented in this contribution aims to elucidate the degradation mechanisms of organophosphate pesticides as organic pollutants by X-ray irradiation. Diazinon, one of the most widely used pesticides in Slovenia, has been chosen as a model compound for the study of radiation induced degradation of organophosphorus pesticides. Recent studies have shown that the degradation of the pesticide at normal conditions (room temperature, exposure to daylight) is strongly enhanced when the pesticide is exposed to UV light (Hg lamp, λ=254 nm, XeCl excimer laser, λ=308 nm). In our study we irradiated the pesticide in aqueous media with a white x-ray beam from a conventional x-ray source. Mo X-ray tube operating at the high voltage of 55 kV and a current of 45 mA was used. The flux of the continuous X-ray beam was stabilized within 1%. Saturated water solution of the pesticide (volume: 4mL, conc. of 40 mg L -1 ) was inserted in 1 cm long lucite cell with 1 mm thick lucite windows. The whole volume of the solution in the cell was exposed to the unfiltered X-ray beam. The dose rate on the sample was about 1 mGy/s. Different irradiation times between 30 min to 120 min were chosen to study the dependence of the pesticide decomposition with the absorbed dose. Solid phase extraction was employed for sample extraction from the solution, and gas chromatography was used for the identification and quantification of the compounds. The results show that the concentration of the pesticide in the solution decreases exponentially with the exposure time, i.e. with the absorbed dose. At irradiation conditions described above, the time constant of the exponential decrease was 74 min

  14. White beam synchrotron x-ray topography of gallium arsenide

    International Nuclear Information System (INIS)

    Winter, J.M. Jr.; Green, R.E. Jr.; Corak, W.S.

    1988-01-01

    The defect structure of gallium arsenide was investigated using white beam transmission topography. The samples were cut and polished monocrystal substrates from different suppliers. The goal of the work was to determine the viability of the method for documenting various crystallographic defect structures and establishing their effect on the performance of integrated microwave circuits fabricated on the wafers. The principles of the technique, essentially identical to classical Laue x-ray diffraction, are outlined. Two distinct defect structures were determined in the topographs. Reasons for the defect structures were postulated and the application of the method for quality control assessments of manufacturer-supplied gallium arsenide substrates was assessed

  15. Electron cyclotron resonance ion source plasma characterization by X-ray spectroscopy and X-ray imaging

    Energy Technology Data Exchange (ETDEWEB)

    Mascali, David, E-mail: davidmascali@lns.infn.it; Castro, Giuseppe; Celona, Luigi; Neri, Lorenzo; Gammino, Santo [INFN–Laboratori Nazionali del Sud, Via S. Sofia 62, 95125 Catania (Italy); Biri, Sándor; Rácz, Richárd; Pálinkás, József [Institute for Nuclear Research (Atomki), Hungarian Academy of Sciences, Bem tér 18/c, H-4026 Debrecen (Hungary); Caliri, Claudia [INFN–Laboratori Nazionali del Sud, Via S. Sofia 62, 95125 Catania (Italy); Università degli Studi di Catania, Dip.to di Fisica e Astronomia, via Santa Sofia 64, 95123 Catania (Italy); Romano, Francesco Paolo [INFN–Laboratori Nazionali del Sud, Via S. Sofia 62, 95125 Catania (Italy); CNR, Istituto per i Beni Archeologici e Monumentali, Via Biblioteca 4, 95124 Catania (Italy); Torrisi, Giuseppe [INFN–Laboratori Nazionali del Sud, Via S. Sofia 62, 95125 Catania (Italy); Università Mediterranea di Reggio Calabria, DIIES, Via Graziella, I-89100 Reggio Calabria (Italy)

    2016-02-15

    An experimental campaign aiming to investigate electron cyclotron resonance (ECR) plasma X-ray emission has been recently carried out at the ECRISs—Electron Cyclotron Resonance Ion Sources laboratory of Atomki based on a collaboration between the Debrecen and Catania ECR teams. In a first series, the X-ray spectroscopy was performed through silicon drift detectors and high purity germanium detectors, characterizing the volumetric plasma emission. The on-purpose developed collimation system was suitable for direct plasma density evaluation, performed “on-line” during beam extraction and charge state distribution characterization. A campaign for correlating the plasma density and temperature with the output charge states and the beam intensity for different pumping wave frequencies, different magnetic field profiles, and single-gas/gas-mixing configurations was carried out. The results reveal a surprisingly very good agreement between warm-electron density fluctuations, output beam currents, and the calculated electromagnetic modal density of the plasma chamber. A charge-coupled device camera coupled to a small pin-hole allowing X-ray imaging was installed and numerous X-ray photos were taken in order to study the peculiarities of the ECRIS plasma structure.

  16. EFFECTS OF X-RAY BEAM ANGLE AND GEOMETRIC DISTORTION ON WIDTH OF EQUINE THORACOLUMBAR INTERSPINOUS SPACES USING RADIOGRAPHY AND COMPUTED TOMOGRAPHY

    DEFF Research Database (Denmark)

    Djernaes, Julie D.; Nielsen, Jon V.; Berg, Lise C.

    2017-01-01

    The widths of spaces between the thoracolumbar processi spinosi (interspinous spaces) are frequently assessed using radiography in sports horses; however effects of varying X-ray beam angles and geometric distortion have not been previously described. The aim of this prospective, observational...... study was to determine whether X-ray beam angle has an effect on apparent widths of interspinous spaces. Thoracolumbar spine specimens were collected from six equine cadavers and left-right lateral radiographs and sagittal and dorsal reconstructed computed tomographic (CT) images were acquired...... measurements. Effect of geometric distortion was evaluated by comparing the interspinous space in radiographs with sagittal and dorsal reconstructed CT images. A total of 49 interspinous spaces were sampled, yielding 274 measurements. X-ray beam angle significantly affected measured width of interspinous...

  17. Characteristic parameters analysis on diagnostic X-ray beams for dosemeter calibration

    International Nuclear Information System (INIS)

    Oliveira, Paulo Marcio Campos de

    2008-01-01

    Ionizing radiation metrology is the base to achieve reliable dose measurements in ali areas; it is also part of the framework that is established to assure radiation protection procedures in order to avoid or minimize the harmful biological effect that may be caused by ionizing radiation. A well done metrology means the use of reliable instruments that comply with standard performance requirements worldwide accepted. Those instruments are expected to be calibrated by Metrology Laboratories under well defined conditions. The International Electrotechnical Commission (IEC) in Standard 61267 established the reference radiations for medical diagnostic x-ray equipment that are recommended to be used for calibrating dosimetric systems for diagnostic dosimetry. In this work, X-ray beam qualities were established in a Calibration Laboratory and their characteristics were analyzed through the measurement of beam parameters like inherent tube filtration, beam uniformity and field size, energy spectra and peak voltage for additional filtration with 94.425 por cent and 99.999 por cent purity filters. Also, the first half-value layer and the homogeneity coefficient were measured for the three RQR 2, RQR 6 and RQR 10 IEC beam qualities and they were analyzed according to the IEC standard. Air-kerma measurements were carried out with an ionization chamber that had its reliability confirmed through repetition and reproducibility reading tests. In 50 sets of measurements the maximum standard deviation found of 10 successive readings was 0.19 %; the maximum shift of the reading mean value at a fixed geometry condition was 0.80 % with an overall standard deviation of 0.23 %. Results showed that the use of different purity filters did not cause a relevant influence on the beam energy spectra. An ionization chamber was also calibrated against a standard dosimeter in ali implemented reference radiations and the relevant sources of uncertainties were estimated. Calibration could be done

  18. Scintillating ribbon x-ray detector

    International Nuclear Information System (INIS)

    Kinchen, B.E.; Rogers, A.

    1995-01-01

    A patent in the early 1970's by Aerojet Corporation in Sacramento, CA put forth the idea of using an array of scintillating fibers for x-ray detection and imaging. In about 1975, Pratt and Whitney Aircraft in East Hartford, CT designed and manufactured an imaging system based on the patent. The device was 1.75 in thick in the direction of the x-ray beam and about 4 in. by 4 in. square. The device was used with a 8 MeV x-ray source to image and measure internal clearances within operating aircraft, gas turbines engines. There are significant advantages of fiber optic detectors in x-ray detection. However, the advantages are often outweighed by the disadvantages. Two of the advantages of scintillating fiber optic x-ray detectors are: (1) high limiting spatial frequency -- between 20 and 25 lp/mm; and (2) excellent x-ray stopping power -- they can be made thick and retain spatial resolution. In traditional fiber optic detectors the x-rays are oriented parallel to the long axis of the fiber. For the scintillating ribbon x-ray sensor, the x-rays are oriented normal to the fiber long axis. This ribbon sensor technique has a number of advantages over the two current radiographic techniques digital x-radiography and x-ray film: The main advantage the ribbon has is size and shape. It can be as thin as 0.05 in., virtually any width or length, and flexible. Once positioned in a given location, 20 to 100 square inches of the object being inspected can be imaged with a single x-ray beam sweep. It is clear that conventional digital cameras do not lend themselves to placement between walls of aircraft structures or similar items requiring x-ray inspections. A prototype scintillating ribbon x-ray sensor has been fabricated and tested by Synergistic Detector Designs. Images were acquired on corrosion test panels of aluminum fabricated by Iowa State University

  19. Guides for intraoral x-rays

    International Nuclear Information System (INIS)

    Ogunsunlade, O.A.

    1988-01-01

    An h-shaped exterior guide for use in combination with a SNAP-A-RAY film holder for accurately aligning a beam from an X-ray cone with an X-ray film during the process of taking intraoral periapical dental X-rays of the maxillary and mandibular teeth is described comprising: a first guide arm laterally and detachably connectable through a housing means; a traverse arm extending from the midpoint of the first guide arm and parallel to the X-ray film; and a second guide arm extending perpendicularly from an end of the traverse arm toward a plane of the X-ray film and in parallel relation up to an end point of the first guide arm

  20. Impact of polychromatic x-ray sources on helical, cone-beam computed tomography and dual-energy methods

    International Nuclear Information System (INIS)

    Sidky, Emil Y; Zou Yu; Pan Xiaochuan

    2004-01-01

    Recently, there has been much work devoted to developing accurate and efficient algorithms for image reconstruction in helical, cone-beam computed tomography (CT). Little attention, however, has been directed to the effect of physical factors on helical, cone-beam CT image reconstruction. This work investigates the effect of polychromatic x-rays on image reconstruction in helical, cone-beam computed tomography. A pre-reconstruction dual-energy technique is developed to reduce beam-hardening artefacts and enhance contrast in soft tissue

  1. X-ray phase contrast imaging at MAMI

    International Nuclear Information System (INIS)

    El-Ghazaly, M.; Backe, H.; Lauth, W.; Kube, G.; Kunz, P.; Sharafutdinov, A.; Weber, T.

    2006-01-01

    Experiments have been performed to explore the potential of the low emittance 855 MeV electron beam of the Mainz Microtron MAMI for imaging with coherent X-rays. Transition radiation from a micro-focused electron beam traversing a foil stack served as X-ray source with good transverse coherence. Refraction contrast radiographs of low absorbing materials, in particular polymer strings with diameters between 30 and 450 μm, were taken with a polychromatic transition radiation X-ray source with a spectral distribution in the energy range between 8 and about 40 keV. The electron beam spot size had standard deviation σ h =(8.6±0.1) μm in the horizontal and σ v =(7.5±0.1) μm in the vertical direction. X-ray films were used as detectors. The source-to-detector distance amounted to 11.4 m. The objects were placed in a distance of up to 6m from the X-ray film. Holograms of strings were taken with a beam spot size σ v =(0.50±0.05) μm in vertical direction, and a monochromatic X-ray beam of 6keV energy. A good longitudinal coherence has been obtained by the (111) reflection of a flat silicon single crystal in Bragg geometry. It has been demonstrated that a direct exposure CCD chip with a pixel size of 13 x 13 μm 2 provides a highly efficient on-line detector. Contrast images can easily be generated with a complete elimination of all parasitic background. The on-line capability allows a minimization of the beam spot size by observing the smallest visible interference fringe spacings or the number of visible fringes. It has been demonstrated that X-ray films are also very useful detectors. The main advantage in comparison with the direct exposure CCD chip is the resolution. For the Structurix D3 (Agfa) X-ray film the standard deviation of the resolution was measured to be σ f =(1.2±0.4) μm, which is about a factor of 6 better than for the direct exposure CCD chip. With the small effective X-ray spot size in vertical direction of σ v =(1.2±0.3)μm and a

  2. X-ray phase contrast imaging at MAMI

    Energy Technology Data Exchange (ETDEWEB)

    El-Ghazaly, M.; Backe, H.; Lauth, W.; Kube, G.; Kunz, P.; Sharafutdinov, A.; Weber, T. [Universitaet Mainz, Institut fuer Kernphysik, Mainz (Germany)

    2006-05-15

    Experiments have been performed to explore the potential of the low emittance 855 MeV electron beam of the Mainz Microtron MAMI for imaging with coherent X-rays. Transition radiation from a micro-focused electron beam traversing a foil stack served as X-ray source with good transverse coherence. Refraction contrast radiographs of low absorbing materials, in particular polymer strings with diameters between 30 and 450 {mu}m, were taken with a polychromatic transition radiation X-ray source with a spectral distribution in the energy range between 8 and about 40 keV. The electron beam spot size had standard deviation {sigma}{sub h}=(8.6{+-}0.1) {mu}m in the horizontal and {sigma}{sub v}=(7.5{+-}0.1) {mu}m in the vertical direction. X-ray films were used as detectors. The source-to-detector distance amounted to 11.4 m. The objects were placed in a distance of up to 6m from the X-ray film. Holograms of strings were taken with a beam spot size {sigma}{sub v}=(0.50{+-}0.05) {mu}m in vertical direction, and a monochromatic X-ray beam of 6keV energy. A good longitudinal coherence has been obtained by the (111) reflection of a flat silicon single crystal in Bragg geometry. It has been demonstrated that a direct exposure CCD chip with a pixel size of 13 x 13 {mu}m{sup 2} provides a highly efficient on-line detector. Contrast images can easily be generated with a complete elimination of all parasitic background. The on-line capability allows a minimization of the beam spot size by observing the smallest visible interference fringe spacings or the number of visible fringes. It has been demonstrated that X-ray films are also very useful detectors. The main advantage in comparison with the direct exposure CCD chip is the resolution. For the Structurix D3 (Agfa) X-ray film the standard deviation of the resolution was measured to be {sigma}{sub f}=(1.2{+-}0.4) {mu}m, which is about a factor of 6 better than for the direct exposure CCD chip. With the small effective X-ray spot size

  3. The Columbia University proton-induced soft x-ray microbeam.

    Science.gov (United States)

    Harken, Andrew D; Randers-Pehrson, Gerhard; Johnson, Gary W; Brenner, David J

    2011-09-15

    A soft x-ray microbeam using proton-induced x-ray emission (PIXE) of characteristic titanium (K(α) 4.5 keV) as the x-ray source has been developed at the Radiological Research Accelerator Facility (RARAF) at Columbia University. The proton beam is focused to a 120 μm × 50 μm spot on the titanium target using an electrostatic quadrupole quadruplet previously used for the charged particle microbeam studies at RARAF. The proton induced x-rays from this spot project a 50 μm round x-ray generation spot into the vertical direction. The x-rays are focused to a spot size of 5 μm in diameter using a Fresnel zone plate. The x-rays have an attenuation length of (1/e length of ~145 μm) allowing more consistent dose delivery across the depth of a single cell layer and penetration into tissue samples than previous ultra soft x-ray systems. The irradiation end station is based on our previous design to allow quick comparison to charged particle experiments and for mixed irradiation experiments.

  4. X-ray diffraction imaging with the Multiple Inverse Fan Beam topology: Principles, performance and potential for security screening

    Energy Technology Data Exchange (ETDEWEB)

    Harding, G., E-mail: Geoffrey.Harding@Morphodetection.com [Morpho Detection Germany GmbH, Heselstuecken 3, 22453 Hamburg (Germany); Fleckenstein, H.; Kosciesza, D.; Olesinski, S.; Strecker, H.; Theedt, T.; Zienert, G. [Morpho Detection Germany GmbH, Heselstuecken 3, 22453 Hamburg (Germany)

    2012-07-15

    The steadily increasing number of explosive threat classes, including home-made explosives (HMEs), liquids, amorphous and gels (LAGs), is forcing up the false-alarm rates of security screening equipment. This development can best be countered by increasing the number of features available for classification. X-ray diffraction intrinsically offers multiple features for both solid and LAGs explosive detection, and is thus becoming increasingly important for false-alarm and cost reduction in both carry-on and checked baggage security screening. Following a brief introduction to X-ray diffraction imaging (XDI), which synthesizes in a single modality the image-forming and material-analysis capabilities of X-rays, the Multiple Inverse Fan Beam (MIFB) XDI topology is described. Physical relationships obtaining in such MIFB XDI components as the radiation source, collimators and room-temperature detectors are presented with experimental performances that have been achieved. Representative X-ray diffraction profiles of threat substances measured with a laboratory MIFB XDI system are displayed. The performance of Next-Generation (MIFB) XDI relative to that of the 2nd Generation XRD 3500{sup TM} screener (Morpho Detection Germany GmbH) is assessed. The potential of MIFB XDI, both for reducing the exorbitant cost of false alarms in hold baggage screening (HBS), as well as for combining 'in situ' liquid and solid explosive detection in carry-on luggage screening is outlined. - Highlights: Black-Right-Pointing-Pointer X-ray diffraction imaging (XDI) synthesizes analysis and imaging in one x-ray modality. Black-Right-Pointing-Pointer A novel XDI beam topology comprising multiple inverse fan-beams (MIFB) is described. Black-Right-Pointing-Pointer The MIFB topology is technically easy to realize and has high photon collection efficiency. Black-Right-Pointing-Pointer Applications are envisaged in checkpoint, hold baggage and cargo screening.

  5. New imaging technique based on diffraction of a focused x-ray beam

    Energy Technology Data Exchange (ETDEWEB)

    Kazimirov, A [Cornell High Energy Synchrotron Source (CHESS), Cornell University, Ithaca, NY 14853 (United States); Kohn, V G [Russian Research Center ' Kurchatov Institute, 123182 Moscow (Russian Federation); Cai, Z-H [Advanced Photon Source, 9700 S. Cass Avenue, Argonne, IL 60439 (United States)], E-mail: ayk7@cornell.edu

    2009-01-07

    We present first experimental results from a new diffraction depth-sensitive imaging technique. It is based on the diffraction of a focused x-ray beam from a crystalline sample and recording the intensity pattern on a high-resolution CCD detector positioned at a focal plane. Structural non-uniformity inside the sample results in a region of enhanced intensity in the diffraction pattern. The technique was applied to study silicon-on-insulator thin layers of various thicknesses which revealed a complex strain profile within the layers. A circular Fresnel zone plate was used as a focusing optic. Incoherent diffuse scattering spreads out of the diffraction plane and results in intensity recorded outside of the focal spot providing a new approach to separately register x-rays scattered coherently and incoherently from the sample. (fast track communication)

  6. E-line: A new crystal collimator beam line for source size measurements at CHESS

    Energy Technology Data Exchange (ETDEWEB)

    White, Jeffrey A. [CHESS, Cornell High Energy Synchrotron Source, Cornell University, Ithaca, NY 14850-8001 (United States)], E-mail: jaw7@cornell.edu; Revesz, Peter; Finkelstein, Ken [CHESS, Cornell High Energy Synchrotron Source, Cornell University, Ithaca, NY 14850-8001 (United States)

    2007-11-11

    A new X-ray beam line has been constructed at cornell high energy synchrotron source (CHESS) to measure the vertical and horizontal source size of the positron particle beam. The cornell laboratory of elementary particle physics (LEPP) operates the storage ring (CESR) for X-ray generation for the CHESS user community by circulating electrons and their antimatter counterpart positrons in counter-rotating beams. As the laboratory reduces the emittances of particle beams to increase X-ray brilliance, there has been an increasing need for diagnostic tools to measure and monitor source size. A beam line front end that accesses the positron synchrotron light has been fitted with an experimental chamber and apparatus of compact design capable of horizontal and vertical source size measurement using the 'crystal collimator' technique, and an additional setup for vertical beam position monitoring using a luminescence-based X-ray video beam position monitoring system. The crystal collimators each consist of two Si(2 2 0) crystals in a dispersive (+,+) arrangement that diffract X-rays to a fluorescent material coated on a view port observed with a CCD camera. Measurements of the positron vertical beam size using the crystal collimation method at E-line are compared with measurements of visible synchrotron light at a remotely located dedicated port on the storage ring.

  7. Evaluation of the spectral distribution of X-ray beams from measurements on the scattered radiation

    International Nuclear Information System (INIS)

    Casnati, E.; Baraldi, C.

    1980-01-01

    Most of the phenomena activated by photons with energies below 100 keV show an apparent or real dependence on the quantum energy. Therefore, knowledge of the beam energy characteristics is of primary importance for interpretation of the irradiation results. The greatest difficulty arises from the high flux density of the beams usually employed which does not allow direct measurements of the beam. A method was developed which permits evaluation of the spectral distribution of the X-ray beam from a spectrometric measurement of the radiation scattered by a thin foil of a suitable metal. This makes possible a new and more rational approach to the measurement of X-rays in the energy range where the interaction parameters show a large photon energy dependence. The corrections required by the presence of some collateral effects, among which the most important is the coexistence of the coherent and incoherent scattering, must be evaluated. The knowledge of the spectral distribution is of immediate usefulness for studies of radiation damage in biological and other materials, for the calibration of radiation measuring instruments and for the improvement of the radiological instrumentation response which contributes to reducing the patient's dose. (H.K.)

  8. Nondestructive Detection of the Internalquality of Apple Using X-Ray and Machine Vision

    Science.gov (United States)

    Yang, Fuzeng; Yang, Liangliang; Yang, Qing; Kang, Likui

    The internal quality of apple is impossible to be detected by eyes in the procedure of sorting, which could reduce the apple’s quality reaching market. This paper illustrates an instrument using X-ray and machine vision. The following steps were introduced to process the X-ray image in order to determine the mould core apple. Firstly, lifting wavelet transform was used to get a low frequency image and three high frequency images. Secondly, we enhanced the low frequency image through image’s histogram equalization. Then, the edge of each apple's image was detected using canny operator. Finally, a threshold was set to clarify mould core and normal apple according to the different length of the apple core’s diameter. The experimental results show that this method could on-line detect the mould core apple with less time consuming, less than 0.03 seconds per apple, and the accuracy could reach 92%.

  9. Dose reduction by x-ray beam filtration in screen-film radiography

    International Nuclear Information System (INIS)

    Koedooder, C.

    1986-01-01

    This thesis describes experimental and theoretical aspects of dose reduction by x-ray beam filtration in screen-film radiography. The thesis deals mainly with dose reduction under the constraint of constant image quality; an analytical approach is chosen. Therefore, part of the thesis deals with the development of an algorithm to calculate patient dose and exposure for different filter materials and different tube load conditions, under the constraint of constant contrast and constant optical density. (Auth.)

  10. Application of a tandem ionization chamber in a quality control program of X-ray beams, radiotherapy level

    International Nuclear Information System (INIS)

    Yoshizumi, Maira T.; Caldas, Linda V.E.

    2008-01-01

    A tandem ionization chamber, developed at the Instituto de Pesquisas Energeticas e Nucleares (IPEN), for X radiation beams, radiotherapy level, was applied into a quality control program of the Calibration Laboratory of IPEN. This ionization chamber is composed by two ionization chambers, with a volume of 0.6 cm 3 each one. Its inner plane-parallel electrodes and guard rings are made of different materials: one is made of aluminum and the other is made of graphite. Because of this difference in materials, the ionization chamber forms a tandem system. The relative response of the calibration factors of both sides of the chamber allows an easy verification of the X-ray beam qualities stability. The ionization chamber was submitted to some tests to verify the stability of its response: leakage current before and after exposure, repeatability and reproducibility. The performance of the ionization chamber was satisfactory. (author)

  11. Characterization of Polycrystalline Materials Using Synchrotron X-ray Imaging and Diffraction Techniques

    DEFF Research Database (Denmark)

    Ludwig, Wolfgang; King, A.; Herbig, M.

    2010-01-01

    The combination of synchrotron radiation x-ray imaging and diffraction techniques offers new possibilities for in-situ observation of deformation and damage mechanisms in the bulk of polycrystalline materials. Minute changes in electron density (i.e., cracks, porosities) can be detected using...... propagation based phase contrast imaging, a 3-D imaging mode exploiting the coherence properties of third generation synchrotron beams. Furthermore, for some classes of polycrystalline materials, one may use a 3-D variant of x-ray diffraction imaging, termed x-ray diffraction contrast tomography. X-ray...

  12. Measure of the attenuation curve of a beam of X-rays with TLD-100 dosimeters of LiF

    International Nuclear Information System (INIS)

    Bonzi, E. V.; Mainardi, R. T.; Germanier, A.; Delgado, V.

    2011-10-01

    The attenuation curve of a beam of X-rays represents the beam intensity in function of the attenuator thickness interposed between the source and the detector. To know with the major possible precision the attenuation curve is indispensable in procedures of spectral reconstruction. Their periodic measuring also offers valuable information on the correct operation of a tube of X-rays, diagnostic or therapy, when not have a specific detector for that activity. In this work was measured the attenuation curve of a tube of X-rays operated to 50 kV and 0.5 ma, using existent elements in any diagnostic or therapy laboratory with radiations. In the measures commercial aluminum foil was used, bent until 24 times and thermoluminescent dosimeters TLD 100 - LiF. Also, for comparison, was measured this attenuation curve with an ionization chamber brand Capintec model 192. Was determined by X-rays fluorescence the composition of the aluminium foil, since the present elements in the alloy can to affect the form of the attenuation curve. It is interesting to observe that these elements are in very low proportion (ppm) that they do not alter the attenuation capacity of the pure aluminium. Finally in a precision balance we weigh a big piece (30 cm x 100 cm) of aluminium foil and we obtained the thickness in g/c m2. It is possible to obtain attenuation curves of a beam of X-rays, with a high precision procedure and reproducibility. The use of TLD-100 dosimeters of LiF or similar makes that this activity was also quick and simple. (Author)

  13. Machine recognition of navel orange worm damage in X-ray images of pistachio nuts

    Energy Technology Data Exchange (ETDEWEB)

    Keagy, P.M.; Schatzki, T.F. [USDA-ARS Western Regional Research Center, Albany, CA (United States); Parvin, B. [Lawrence Berkeley Lab., CA (United States)

    1994-11-01

    Insect infestation increases the probability of aflatoxin contamination in pistachio nuts. A non-destructive test is currently not available to determine the insect content of pistachio nuts. This paper presents the use of film X-ray images of various types of pistachio nuts to assess the possibility of machine recognition of insect infested nuts. Histogram parameters of four derived images are used in discriminant functions to select insect infested nuts from specific processing streams.

  14. The development and operation of a method for the remote determination of X-ray beam parameters used in dental radiography

    International Nuclear Information System (INIS)

    Hewitt, J.M.

    1984-07-01

    The method described is a part of the Dental Monitoring Service operated by the Board in the UK for the assessment of radiation protection in dental practice. This postal service, which provides a comprehensive survey of dental X-ray sets and radiographic procedures, is undertaken by means of a questionnaire, film cassettes for exposure to the X-ray set and a personal monitoring component to check operator doses. The film cassettes and the methods by which the X-ray beam parameters are obtained are described in detail. The cassettes use radiation monitoring film to realise, by means of measurements of relative transmission through selected copper filters, the extended dynamic range of exposure necessary for accurate indication of the operating kilovoltage and total beam filtration. The standard of the X-ray unit with regard to the relevant regulations and code of practice can then be assessed, and, from the values of radiation dose determined for chosen exposure times, exposure settings for optimum quality radiographs can be recommended where appropriate. Although designed primarily for dental X-ray units, use of the film cassette package may be extended, with suitable calibration, to general diagnostic X-ray survey measurements. (author)

  15. X-ray study of a test quadrant of the SODART telescopes using the expanded beam x-ray optics facility at the Daresbury synchrotron

    DEFF Research Database (Denmark)

    Christensen, Finn Erland; Hornstrup, Allan; Frederiksen, P.

    1994-01-01

    The imaging properties of a test model of the SODART telescopes have been studied using an expanded beam X-ray facility at the Daresbury synchrotron. The encircled power and the point spread function at three energies 6.627 keV, 8.837 keV and 11.046 keV have been measured using 1D and 2D position...

  16. New x-ray parallel beam facility XPBF 2.0 for the characterization of silicon pore optics

    Science.gov (United States)

    Krumrey, Michael; Müller, Peter; Cibik, Levent; Collon, Max; Barrière, Nicolas; Vacanti, Giuseppe; Bavdaz, Marcos; Wille, Eric

    2016-07-01

    A new X-ray parallel beam facility (XPBF 2.0) has been installed in the laboratory of the Physikalisch-Technische Bundesanstalt at the synchrotron radiation facility BESSY II in Berlin to characterize silicon pore optics (SPOs) for the future X-ray observatory ATHENA. As the existing XPBF which is operated since 2005, the new beamline provides a pencil beam of very low divergence, a vacuum chamber with a hexapod system for accurate positioning of the SPO to be investigated, and a vertically movable CCD-based camera system to register the direct and the reflected beam. In contrast to the existing beamline, a multilayer-coated toroidal mirror is used for beam monochromatization at 1.6 keV and collimation, enabling the use of beam sizes between about 100 μm and at least 5 mm. Thus the quality of individual pores as well as the focusing properties of large groups of pores can be investigated. The new beamline also features increased travel ranges for the hexapod to cope with larger SPOs and a sample to detector distance of 12 m corresponding to the envisaged focal length of ATHENA.

  17. Computer simulation of the CSPAD, ePix10k, and RayonixMX170HS X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Tina, Adrienne

    2015-08-21

    The invention of free-electron lasers (FELs) has opened a door to an entirely new level of scientific research. The Linac Coherent Light Source (LCLS) at SLAC National Accelerator Laboratory is an X-ray FEL that houses several instruments, each with its own unique X-ray applications. This light source is revolutionary in that while its properties allow for a whole new range of scientific opportunities, it also poses numerous challenges. For example, the intensity of a focused X-ray beam is enough to damage a sample in one mere pulse; however, the pulse speed and extreme brightness of the source together are enough to obtain enough information about that sample, so that no further measurements are necessary. An important device in the radiation detection process, particularly for X-ray imaging, is the detector. The power of the LCLS X-rays has instigated a need for better performing detectors. The research conducted for this project consisted of the study of X-ray detectors to imitate their behaviors in a computer program. The analysis of the Rayonix MX170-HS, CSPAD, and ePix10k in particular helped to understand their properties. This program simulated the interaction of X-ray photons with these detectors to discern the patterns of their responses. A scientist’s selection process of a detector for a specific experiment is simplified from the characterization of the detectors in the program.

  18. Blade-type X-ray beam position monitors for SPring-8 undulator beamlines

    CERN Document Server

    Aoyagi, H; Kitamura, H

    2001-01-01

    The X-ray beam position monitors had been designed and installed for SPring-8 insertion device beamlines. These monitors are being utilized for photon beam diagnostics. The beam from the standard undulator in SPring-8 has the total power of 11 kW and the power density of 470 kW/mrad sup 2 , typically. Each monitor has four CVD diamond blades coated with metal for detector heads. We have already introduced three styles of monitors to match various insertion devices in SPring-8. A standard style, or a fixed-blade style, is used mainly for a standard in-vacuum undulator beamlines. A horizontal-blade-drive style and a four-blade-drive style are used for beamlines of a wiggler and a twin helical undulator that have wide power distributions, and for figure-8 undulators that have asymmetric power distributions, respectively. This report describes the design and the structure of these monitors and the beam-tests for the photon beam diagnostics in detail.

  19. A method for measuring the time structure of synchrotron x-ray beams

    International Nuclear Information System (INIS)

    Moses, W.W.; Derenzo, S.E.

    1989-08-01

    We describe a method employing a plastic scintillator coupled to a fast photomultiplier tube to generate a timing pulse from the x-ray bursts emitted from a synchrotron radiation source. This technique is useful for performing synchrotron experiments where detailed knowledge of the timing distribution is necessary, such as time resolved spectroscopy or fluorescence lifetime experiments. By digitizing the time difference between the timing signal generated on one beam crossing with the timing signal generated on the next beam crossing, the time structure of a synchrotron beam can be analyzed. Using this technique, we have investigated the single bunch time structure at the National Synchrotron Light Source (NSLS) during pilot runs in January, 1989, and found that the majority of the beam (96%) is contained in one rf bucket, while the remainder of the beam (4%) is contained in satellite rf buckets preceeding and following the main rf bucket by 19 ns. 1 ref., 4 figs

  20. X-ray phase contrast imaging at MAMI

    Science.gov (United States)

    El-Ghazaly, M.; Backe, H.; Lauth, W.; Kube, G.; Kunz, P.; Sharafutdinov, A.; Weber, T.

    2006-05-01

    Experiments have been performed to explore the potential of the low emittance 855MeV electron beam of the Mainz Microtron MAMI for imaging with coherent X-rays. Transition radiation from a micro-focused electron beam traversing a foil stack served as X-ray source with good transverse coherence. Refraction contrast radiographs of low absorbing materials, in particular polymer strings with diameters between 30 and 450μm, were taken with a polychromatic transition radiation X-ray source with a spectral distribution in the energy range between 8 and about 40keV. The electron beam spot size had standard deviation σh = (8.6±0.1)μm in the horizontal and σv = (7.5±0.1)μm in the vertical direction. X-ray films were used as detectors. The source-to-detector distance amounted to 11.4m. The objects were placed in a distance of up to 6m from the X-ray film. Holograms of strings were taken with a beam spot size σv = (0.50±0.05)μm in vertical direction, and a monochromatic X-ray beam of 6keV energy. A good longitudinal coherence has been obtained by the (111) reflection of a flat silicon single crystal in Bragg geometry. It has been demonstrated that a direct exposure CCD chip with a pixel size of 13×13μm^2 provides a highly efficient on-line detector. Contrast images can easily be generated with a complete elimination of all parasitic background. The on-line capability allows a minimization of the beam spot size by observing the smallest visible interference fringe spacings or the number of visible fringes. It has been demonstrated that X-ray films are also very useful detectors. The main advantage in comparison with the direct exposure CCD chip is the resolution. For the Structurix D3 (Agfa) X-ray film the standard deviation of the resolution was measured to be σf = (1.2±0.4)μm, which is about a factor of 6 better than for the direct exposure CCD chip. With the small effective X-ray spot size in vertical direction of σv = (1.2±0.3)μm and a geometrical

  1. Compact X-ray source based on Compton backscattering

    CERN Document Server

    Bulyak, E V; Zelinsky, A; Karnaukhov, I; Kononenko, S; Lapshin, V G; Mytsykov, A; Telegin, Yu P; Khodyachikh, A; Shcherbakov, A; Molodkin, V; Nemoshkalenko, V; Shpak, A

    2002-01-01

    The feasibility study of an intense X-ray source based on the interaction between the electron beam in a compact storage ring and the laser pulse accumulated in an optical resonator is carried out. We propose to reconstruct the 160 MeV electron storage ring N-100, which was shutdown several years ago. A new magnetic lattice will provide a transverse of electron beam size of approx 35 mu m at the point of electron beam-laser beam interaction. The proposed facility is to generate X-ray beams of intensity approx 2.6x10 sup 1 sup 4 s sup - sup 1 and spectral brightness approx 10 sup 1 sup 2 phot/0.1%bw/s/mm sup 2 /mrad sup 2 in the energy range from 10 keV up to 0.5 MeV. These X-ray beam parameters meet the requirements for most of technological and scientific applications. Besides, we plan to use the new facility for studying the laser cooling effect.

  2. Compact X-ray source based on Compton backscattering

    Energy Technology Data Exchange (ETDEWEB)

    Bulyak, E.; Gladkikh, P.; Zelinsky, A. E-mail: zelinsky@kipt.kharkov.ua; Karnaukhov, I.; Kononenko, S.; Lapshin, V.; Mytsykov, A.; Telegin, Yu.; Khodyachikh, A.; Shcherbakov, A.; Molodkin, V.; Nemoshkalenko, V.; Shpak, A

    2002-07-21

    The feasibility study of an intense X-ray source based on the interaction between the electron beam in a compact storage ring and the laser pulse accumulated in an optical resonator is carried out. We propose to reconstruct the 160 MeV electron storage ring N-100, which was shutdown several years ago. A new magnetic lattice will provide a transverse of electron beam size of {approx}35 {mu}m at the point of electron beam-laser beam interaction. The proposed facility is to generate X-ray beams of intensity {approx}2.6x10{sup 14} s{sup -1} and spectral brightness {approx}10{sup 12} phot/0.1%bw/s/mm{sup 2}/mrad{sup 2} in the energy range from 10 keV up to 0.5 MeV. These X-ray beam parameters meet the requirements for most of technological and scientific applications. Besides, we plan to use the new facility for studying the laser cooling effect.

  3. Fabrication of mm-wave undulator cavities using deep x-ray lithography

    International Nuclear Information System (INIS)

    Song, J.J.; Kang, Y.W.; Kustom, R.L.; Lai, B.; Nassiri, A.; Feinerman, A.D.; White, V.; Well, G.M.

    1995-01-01

    The possibility of fabricating mm-wave radio frequency cavities (100-300 GHz) using deep x-ray lithography (DXRL) is being investigated. The fabrication process includes manufacture of precision x-ray masks, exposure of positive resist by x-ray through the mask, resist development, and electroforming of the final microstructure. Highly precise, two-dimensional features can be machined onto wafers using DXRL. Major challenges are: fabrication of the wafers into three-dimensional rf structures; alignment and overlay accuracy of structures; adhesion of the PMMA on the copper substrate; and selection of a developer to obtain high resolution. Rectangular cavity geometry is best suited to this fabrication technique. A 30- or 84-cell 108-GHz mm-wave structure can serve as an electromagnetic undulator. A mm-wave undulator, which will be discussed later, may have special features compared to the conventional undulator. First harmonic undulator radiation at 5.2 KeV would be possible using the Advanced Photon Source (APS) linac system, which provides a low-emittance electron beam by using an rf thermionic gun with an energy as high as 750-MeV. More detailed rf simulation, heat extraction analysis, beam dynamics using a mm-wave structure, and measurements on lOx larger scale models can be found in these proceedings

  4. Automated three-dimensional X-ray analysis using a dual-beam FIB

    International Nuclear Information System (INIS)

    Schaffer, Miroslava; Wagner, Julian; Schaffer, Bernhard; Schmied, Mario; Mulders, Hans

    2007-01-01

    We present a fully automated method for three-dimensional (3D) elemental analysis demonstrated using a ceramic sample of chemistry (Ca)MgTiO x . The specimen is serially sectioned by a focused ion beam (FIB) microscope, and energy-dispersive X-ray spectrometry (EDXS) is used for elemental analysis of each cross-section created. A 3D elemental model is reconstructed from the stack of two-dimensional (2D) data. This work concentrates on issues arising from process automation, the large sample volume of approximately 17x17x10 μm 3 , and the insulating nature of the specimen. A new routine for post-acquisition data correction of different drift effects is demonstrated. Furthermore, it is shown that EDXS data may be erroneous for specimens containing voids, and that back-scattered electron images have to be used to correct for these errors

  5. ADVANCED X-BAND TEST ACCELERATOR FOR HIGH BRIGHTNESS ELECTRON AND GAMMA RAY BEAMS

    Energy Technology Data Exchange (ETDEWEB)

    Marsh, R A; Anderson, S G; Barty, C P; Chu, T S; Ebbers, C A; Gibson, D J; Hartemann, F V; Adolphsen, C; Jongewaard, E N; Raubenheimer, T; Tantawi, S G; Vlieks, A E; Wang, J W

    2010-05-12

    In support of Compton scattering gamma-ray source efforts at LLNL, a multi-bunch test stand is being developed to investigate accelerator optimization for future upgrades. This test stand will enable work to explore the science and technology paths required to boost the current 10 Hz monoenergetic gamma-ray (MEGa-Ray) technology to an effective repetition rate exceeding 1 kHz, potentially increasing the average gamma-ray brightness by two orders of magnitude. Multiple bunches must be of exceedingly high quality to produce narrow-bandwidth gamma-rays. Modeling efforts will be presented, along with plans for a multi-bunch test stand at LLNL. The test stand will consist of a 5.5 cell X-band rf photoinjector, single accelerator section, and beam diagnostics. The photoinjector will be a high gradient standing wave structure, featuring a dual feed racetrack coupler. The accelerator will increase the electron energy so that the emittance can be measured using quadrupole scanning techniques. Multi-bunch diagnostics will be developed so that the beam quality can be measured and compared with theory. Design will be presented with modeling simulations, and layout plans.

  6. Soft x-ray lasers

    International Nuclear Information System (INIS)

    Matthews, D.L.; Rosen, M.D.

    1988-01-01

    One of the elusive dreams of laser physicists has been the development of an x-ray laser. After 25 years of waiting, the x-ray laser has at last entered the scientific scene, although those now in operation are still laboratory prototypes. They produce soft x rays down to about five nanometers. X-ray lasers retain the usual characteristics of their optical counterparts: a very tight beam, spatial and temporal coherence, and extreme brightness. Present x-ray lasers are nearly 100 times brighter that the next most powerful x-ray source in the world: the electron synchrotron. Although Lawrence Livermore National Laboratory (LLNL) is widely known for its hard-x-ray laser program which has potential applications in the Strategic Defense Initiative, the soft x-ray lasers have no direct military applications. These lasers, and the scientific tools that result from their development, may one day have a place in the design and diagnosis of both laser fusion and hard x-ray lasers. The soft x-ray lasers now in operation at the LLNL have shown great promise but are still in the primitive state. Once x-ray lasers become reliable, efficient, and economical, they will have several important applications. Chief among them might be the creation of holograms of microscopic biological structures too small to be investigated with visible light. 5 figs

  7. Synchrotron X-ray micro-beam studies of ancient Egyptian make-up

    Science.gov (United States)

    Martinetto, P.; Anne, M.; Dooryhée, E.; Drakopoulos, M.; Dubus, M.; Salomon, J.; Simionovici, A.; Walter, Ph.

    2001-07-01

    Vases full of make-up are most often present in the burial furniture of Egyptian tombs dated from the pharaonic period. The powdered cosmetics made of isolated grains are analysed to identify their trace element signature. From this signature we identify the provenance of the mineral ingredients in the make-up and we observe different impurities in products, which have been demonstrated as synthetic substances by previous works. Focused X-ray micro-beam ( 2×5 μm2) is successively tuned at 11 keV, below the L III absorption edge of Pb, and 31.8 keV for global characterisation of the metal impurities. The fluorescence signal integrated over each single grain is detected against the X-ray micro-diffraction pattern collected in transmission with a bi-dimensional detector. Furthermore, for galena grains rich in Zn, the XANES signal at the K-absorption edge of Zn shows its immediate nearest-neighbour environment.

  8. Characterisation of microfocused beam for synchrotron powder diffraction using a new X-ray camera

    International Nuclear Information System (INIS)

    Thomas, C; Potter, J; Tang, C C; Lennie, A R

    2012-01-01

    The powder diffraction beamline I11, Diamond Light Source, is being continually upgraded as requirements of the user community evolve. Intensities of X-rays from the I11 in-vacuum electron undulator in the 3 GeV synchrotron fall off at higher energies. By focusing higher energy X-rays, we can overcome flux limitations, and open up new diffraction experiments. Here, we describe characterisation of microfocusing using compound refractive lenses (CRL). For a relatively modest outlay, we have developed an experimental setup and a novel X-ray camera with good sensitivity and a resolution specification suitable for characterising these focusing optics. We show that vertical oscillations in the focused beam compromise resolution of the source imaged by the CRL. Nevertheless, we have measured CRL focusing properties, and demonstrate the use of energy scanning to determine lens alignment. Real benefits of the intensity gain are illustrated.

  9. Field size and centring for conventional X-ray equipment

    International Nuclear Information System (INIS)

    Klimpel, H.; Kreienfeld, H.; Overbeck, R.

    1989-01-01

    Since 1973, all X-ray equipment for medical applications in the Federal Republic of Germany has had to be examined according to the requirements of the German ''Rontgenverordnung'' before it is used on patients and after each essential modification of design or construction. These examinations are carried out by inspectors appointed by the authorities, e.g. TUV. The field size adjustment and the centring of the radiation beam in relation to the image reception area is checked, along with other tests. To increase quality assurance in X-ray diagnosis, since the mid-1980s X-ray equipment has also been subject to in-service inspections to an increasing extent. (author)

  10. Precise material identification method based on a photon counting technique with correction of the beam hardening effect in X-ray spectra

    International Nuclear Information System (INIS)

    Kimoto, Natsumi; Hayashi, Hiroaki; Asahara, Takashi; Mihara, Yoshiki; Kanazawa, Yuki; Yamakawa, Tsutomu; Yamamoto, Shuichiro; Yamasaki, Masashi; Okada, Masahiro

    2017-01-01

    The aim of our study is to develop a novel material identification method based on a photon counting technique, in which the incident and penetrating X-ray spectra are analyzed. Dividing a 40 kV X-ray spectra into two energy regions, the corresponding linear attenuation coefficients are derived. We can identify the materials precisely using the relationship between atomic number and linear attenuation coefficient through the correction of the beam hardening effect of the X-ray spectra. - Highlights: • We propose a precise material identification method to be used as a photon counting system. • Beam hardening correction is important, even when the analysis is applied to the short energy regions in the X-ray spectrum. • Experiments using a single probe-type CdTe detector were performed, and Monte Carlo simulation was also carried out. • We described the applicability of our method for clinical diagnostic X-ray imaging in the near future.

  11. X-ray backscatter imaging with a spiral scanner

    International Nuclear Information System (INIS)

    Bossi, R.H.; Cline, J.L.; Friddell, K.D.

    1989-01-01

    X-ray backscatter imaging allows radiographic inspections to be performed with access to only one side of the object. A collimated beam of radiation striking an object will scatter x-rays by Compton scatter and x-ray fluorescence. A detector located on the source side of the part will measure the backscatter signal. By plotting signal strength as gray scale intensity vs. beam position on the object, an image of the object can be constructed. A novel approach to the motion of the collimated incident beam is a spiral scanner. The spiral scanner approach, described in this paper, can image an area of an object without the synchronized motion of the object or detector, required by other backscatter imaging techniques. X-ray backscatter is particularly useful for flaw detection in light element materials such as composites. The ease of operation and the ability to operate non-contact from one side of an object make x-ray backscatter imaging of increasing interest to industrial inspection problems

  12. X-ray spectroscopy of highly-ionized atoms in an electron beam ion trap (EBIT)

    International Nuclear Information System (INIS)

    Marrs, R.E.; Bennett, C.; Chen, M.H.

    1988-01-01

    An Electron Beam Ion Trap at Lawrence Livermore National Laboratory is being used to produce and trap very-highly-charged-ions (q /le/ 70+) for x-ray spectroscopy measurements. Recent measurements of dielectronic recombination, electron impact excitation and transition energies are presented. 15 refs., 12 figs., 1 tab

  13. Optical and mechanical design of the extended x-ray absorption fine structure (EXAFS) beam-line at Indus-II synchrotron source

    International Nuclear Information System (INIS)

    Das, N.C.; Jha, S.N.; Bhattacharyya, D.; Sinha, A.K.; Mishra, V.K.; Verma, Vishnu; Ghosh, A.K.

    2002-11-01

    An extended x-ray absorption fine structure (EXAFS) beam line for x-ray absorption studies using energy dispersive geometry and position sensitive detector is being designed for the INDUS-II Synchrotron source. The beam line would be used for doing x-ray absorption experiments involving measurements of fme structures above the absorption edge of different species of atoms in a material The results of the above experiments would lead to the determination of different important structural parameters of materials viz.. inter-atomic distance. co-ordination number, degree of disorder and radial distribution function etc. The optical design of the beam line has been completed based on the working principle that a single crystal bent in the shape of an ellipse by a crystal bender would act as a dispersing as well as focusing element. The mechanical design of the beam line including the crystal bender has also been completed and discussed here. Calculations have been done to detennine the temperature profile on the different components of the beam line under exposure to synchrotron radiation and proper cooling channels have been designed to bring down the heat load on the components. (author)

  14. Monte Carlo simulation of x-ray buildup factors of lead and its applications in shielding of diagnostic x-ray facilities

    International Nuclear Information System (INIS)

    Kharrati, Hedi; Agrebi, Amel; Karaoui, Mohamed-Karim

    2007-01-01

    X-ray buildup factors of lead in broad beam geometry for energies from 15 to 150 keV are determined using the general purpose Monte Carlo N-particle radiation transport computer code (MCNP4C). The obtained buildup factors data are fitted to a modified three parameter Archer et al. model for ease in calculating the broad beam transmission with computer at any tube potentials/filters combinations in diagnostic energies range. An example for their use to compute the broad beam transmission at 70, 100, 120, and 140 kVp is given. The calculated broad beam transmission is compared to data derived from literature, presenting good agreement. Therefore, the combination of the buildup factors data as determined and a mathematical model to generate x-ray spectra provide a computationally based solution to broad beam transmission for lead barriers in shielding x-ray facilities

  15. Bright circularly polarized soft X-ray high harmonics for X-ray magnetic circular dichroism.

    Science.gov (United States)

    Fan, Tingting; Grychtol, Patrik; Knut, Ronny; Hernández-García, Carlos; Hickstein, Daniel D; Zusin, Dmitriy; Gentry, Christian; Dollar, Franklin J; Mancuso, Christopher A; Hogle, Craig W; Kfir, Ofer; Legut, Dominik; Carva, Karel; Ellis, Jennifer L; Dorney, Kevin M; Chen, Cong; Shpyrko, Oleg G; Fullerton, Eric E; Cohen, Oren; Oppeneer, Peter M; Milošević, Dejan B; Becker, Andreas; Jaroń-Becker, Agnieszka A; Popmintchev, Tenio; Murnane, Margaret M; Kapteyn, Henry C

    2015-11-17

    We demonstrate, to our knowledge, the first bright circularly polarized high-harmonic beams in the soft X-ray region of the electromagnetic spectrum, and use them to implement X-ray magnetic circular dichroism measurements in a tabletop-scale setup. Using counterrotating circularly polarized laser fields at 1.3 and 0.79 µm, we generate circularly polarized harmonics with photon energies exceeding 160 eV. The harmonic spectra emerge as a sequence of closely spaced pairs of left and right circularly polarized peaks, with energies determined by conservation of energy and spin angular momentum. We explain the single-atom and macroscopic physics by identifying the dominant electron quantum trajectories and optimal phase-matching conditions. The first advanced phase-matched propagation simulations for circularly polarized harmonics reveal the influence of the finite phase-matching temporal window on the spectrum, as well as the unique polarization-shaped attosecond pulse train. Finally, we use, to our knowledge, the first tabletop X-ray magnetic circular dichroism measurements at the N4,5 absorption edges of Gd to validate the high degree of circularity, brightness, and stability of this light source. These results demonstrate the feasibility of manipulating the polarization, spectrum, and temporal shape of high harmonics in the soft X-ray region by manipulating the driving laser waveform.

  16. Optimizing contrast agents with respect to reducing beam hardening in nonmedical X-ray computed tomography experiments.

    Science.gov (United States)

    Nakashima, Yoshito; Nakano, Tsukasa

    2014-01-01

    Iodine is commonly used as a contrast agent in nonmedical science and engineering, for example, to visualize Darcy flow in porous geological media using X-ray computed tomography (CT). Undesirable beam hardening artifacts occur when a polychromatic X-ray source is used, which makes the quantitative analysis of CT images difficult. To optimize the chemistry of a contrast agent in terms of the beam hardening reduction, we performed computer simulations and generated synthetic CT images of a homogeneous cylindrical sand-pack (diameter, 28 or 56 mm; porosity, 39 vol.% saturated with aqueous suspensions of heavy elements assuming the use of a polychromatic medical CT scanner. The degree of cupping derived from the beam hardening was assessed using the reconstructed CT images to find the chemistry of the suspension that induced the least cupping. The results showed that (i) the degree of cupping depended on the position of the K absorption edge of the heavy element relative to peak of the polychromatic incident X-ray spectrum, (ii) (53)I was not an ideal contrast agent because it causes marked cupping, and (iii) a single element much heavier than (53)I ((64)Gd to (79)Au) reduced the cupping artifact significantly, and a four-heavy-element mixture of elements from (64)Gd to (79)Au reduced the artifact most significantly.

  17. Supershort avalanche electron beams and x-ray in high-pressure nanosecond discharges

    International Nuclear Information System (INIS)

    Tarasenko, V F; Baksht, E H; Kostyrya, I D; Lomaev, M I; Rybka, D V

    2008-01-01

    The properties of a supershort avalanche electron beam (S AEB) and X-ray radiation produced using a nanosecond volume discharge are examined. An electron beam of the runaway electrons with amplitude of ∼ 50 A has been obtained in air atmospheric pressure. It is reported that S AEB is formed in the angle above 2π sr. Three groups of the runaway electrons are formed in a gas diode under atmospheric air pressure, when nanosecond voltage pulses with amplitude of hundreds of kilovolts are applied. The electron beam has been generated behind a 45 μm thick AlBe foil in SF 6 and Xe under the pressure of 2 arm, and in He under the pressure of about 12 atm. The paper gives the analysis of a generation mechanism of SAEB.

  18. Images of the laser entrance hole from the static x-ray imager at NIF.

    Science.gov (United States)

    Schneider, M B; Jones, O S; Meezan, N B; Milovich, J L; Town, R P; Alvarez, S S; Beeler, R G; Bradley, D K; Celeste, J R; Dixit, S N; Edwards, M J; Haugh, M J; Kalantar, D H; Kline, J L; Kyrala, G A; Landen, O L; MacGowan, B J; Michel, P; Moody, J D; Oberhelman, S K; Piston, K W; Pivovaroff, M J; Suter, L J; Teruya, A T; Thomas, C A; Vernon, S P; Warrick, A L; Widmann, K; Wood, R D; Young, B K

    2010-10-01

    The static x-ray imager at the National Ignition Facility is a pinhole camera using a CCD detector to obtain images of Hohlraum wall x-ray drive illumination patterns seen through the laser entrance hole (LEH). Carefully chosen filters, combined with the CCD response, allow recording images in the x-ray range of 3-5 keV with 60 μm spatial resolution. The routines used to obtain the apparent size of the backlit LEH and the location and intensity of beam spots are discussed and compared to predictions. A new soft x-ray channel centered at 870 eV (near the x-ray peak of a 300 eV temperature ignition Hohlraum) is discussed.

  19. A multiple CCD X-ray detector and its first operation with synchrotron radiation X-ray beam

    CERN Document Server

    Suzuki, M; Kumasaka, T; Sato, K; Toyokawa, H; Aries, I F; Jerram, P A; Ueki, T

    1999-01-01

    A 4x4 array structure of 16 identical CCD X-ray detector modules, called the multiple CCD X-ray detector system (MCCDX), was submitted to its first synchrotron radiation experiment at the protein crystallography station of the RIKEN beamline (BL45XU) at the SPring-8 facility. An X-ray diffraction pattern of cholesterol powder was specifically taken in order to investigate the overall system performance.

  20. Effect of area x-ray beam equalization on image quality and dose in digital mammography

    Energy Technology Data Exchange (ETDEWEB)

    Wong, Jerry; Xu Tong; Husain, Adeel; Le, Huy; Molloi, Sabee [Department of Radiological Sciences, University of California, Irvine, CA 92697 (United States)

    2004-08-21

    In mammography, thick or dense breast regions persistently suffer from reduced contrast-to-noise ratio (CNR) because of degraded contrast from large scatter intensities and relatively high noise. Area x-ray beam equalization can improve image quality by increasing the x-ray exposure to under-penetrated regions without increasing the exposure to other breast regions. Optimal equalization parameters with respect to image quality and patient dose were determined through computer simulations and validated with experimental observations on a step phantom and an anthropomorphic breast phantom. Three parameters important in equalization digital mammography were considered: attenuator material (Z = 13-92), beam energy (22-34 kVp) and equalization level. A Mo/Mo digital mammography system was used for image acquisition. A prototype 16 x 16 piston driven equalization system was used for preparing patient-specific equalization masks. Simulation studies showed that a molybdenum attenuator and an equalization level of 20 were optimal for improving contrast, CNR and figure of merit (FOM = CNR{sup 2}/dose). Experimental measurements using these parameters showed significant improvements in contrast, CNR and FOM. Moreover, equalized images of a breast phantom showed improved image quality. These results indicate that area beam equalization can improve image quality in digital mammography.

  1. New stereotactic X-ray knife

    International Nuclear Information System (INIS)

    Barish, R.J.; Barish, S.V.

    1988-01-01

    For many years, the irradiation of small volumes of tissue in the brain to necrotizing doses has been investigated as a non-invasive alternative to neurosurgery. We propose a new system in which a precisely machined helmet serves as a multi-port focussed X-ray collimator when it is itself irradiated by a conventional medical linear accelerator run in the electron mode. When the collimator is attached to a stereotactic frame, the geometric accuracy of delivering small radiation fields to the brain is limited primarily by the accuracy of the stereotactic localization, and is relatively independent of the positional stability of the accelerator. Field sizes as small as two millimeters are readily achievable. The problem of low dose rate associated with these small fields is overcome by the use of high electron beam currents

  2. X-ray spectroscopy with EBIT

    International Nuclear Information System (INIS)

    Beiersdorfer, P.; Cauble, R.; Chen, M.; DelGrande, N.; Knapp, D.; Marrs, R.; Osterheld, A.; Reed, K.; Schneider, M.; Scofield, J.; Wong, K.; Vogel, D.; Zasadzinski, R.; Chantrenne, S.; Wargelin, B.

    1992-04-01

    X-ray spectroscopy with the Livermore electron beam ion traps provides data on a wide range of atomic physics issues including ionization, recombination, and excitation cross sections, identification of forbidden transitions, and contributions from relatively and quantum electrodynamics to the transition energies. Here we briefly discuss the source characteristics and x-ray instrumentation, and report measurements of the excitation cross sections of the Kα transitions in heliumlike Ti 20+ as a function of beam energy. The measurements allow detailed comparisons with theoretical predictions of the direct electron-impact excitation cross sections, resonance-excitation contributions, and the electron temperature dependence of the ratio of triplet and singlet lines. The results demonstrate the importance of such measurements for increasing the reliability of x-ray diagnostics of laboratory and astrophysical plasmas

  3. SU-G-201-08: Energy Response of Thermoluminescent Microcube Dosimeters in Water for Kilovoltage X-Ray Beams

    Energy Technology Data Exchange (ETDEWEB)

    Di Maso, L; Lawless, M; Culberson, W; DeWerd, L [University of Wisconsin- Madison, Madison, WI (United States)

    2016-06-15

    Purpose: To characterize the energy dependence for TLD-100 microcubes in water at kilovoltage energies. Methods: TLD-100 microcubes with dimensions of (1 × 1 × 1) mm{sup 3} were irradiated with kilovoltage x-rays in a custom-built thin-window liquid water phantom. The TLD-100 microcubes were held in Virtual Water™ probes and aligned at a 2 cm depth in water. Irradiations were performed using the M-series x-ray beams of energies ranging from 50-250 kVp and normalized to a {sup 60}Co beam located at the UWADCL. Simulations using the EGSnrc Monte Carlo Code System were performed to model the x-ray beams, the {sup 60}Co beam, the water phantom and the dosimeters in the phantom. The egs-chamber user code was used to tally the dose to the TLDs and the dose to water. The measurements and calculations were used to determine the intrinsic energy dependence, absorbed-dose energy dependence, and absorbed-dose sensitivity. These values were compared to TLD-100 chips with dimensions of (3.2 × 0.9 × 0.9) mm{sup 3}. Results: The measured TLD-100 microcube response per dose to water among all investigated x-ray energies had a maximum percent difference of 61% relative to {sup 60}Co. The simulated ratio of dose to water to the dose to TLD had a maximum percent difference of 29% relative to {sup 60}Co. The ratio of dose to TLD to the TLD output had a maximum percent difference of 13% relative to {sup 60}Co. The maximum percent difference for the absorbed-dose sensitivity was 15% more than the used value of 1.41. Conclusion: These results confirm that differences in beam quality have a significant effect on TLD response when irradiated in water. These results also indicated a difference in TLD-100 response between microcube and chip geometries. The intrinsic energy dependence and the absorbed-dose energy dependence deviated up to 10% between TLD-100 microcubes and chips.

  4. U-shape rotating anti-cathode compact X-ray generator: 20 times stronger than the commercially available X-ray source

    Energy Technology Data Exchange (ETDEWEB)

    Sakabe, N., E-mail: sakabe-dsb@sbsp.jp; Sakabe, K. [High Energy Accelerator Research Organization (KEK), 1-1 Oho, Tsukuba 305-0801 (Japan); Foundation for Advancement of International Science (FAIS), Kasuga 3-chome, Tsukuba, Ibaraki 305-0821 (Japan); Ohsawa, S.; Sakai, T.; Kobayakawa, H.; Sugimura, T.; Ikeda, M.; Tawada, M. [High Energy Accelerator Research Organization (KEK), 1-1 Oho, Tsukuba, Ibaraki 305-0801 (Japan); Watanabe, N.; Sasaki, K. [Nagoya University, Chikusa, Nagoya, Aichi 464-8603 (Japan); Wakatsuki, M. [National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Higashi, Tsukuba, Ibaraki 305-8568 (Japan)

    2013-11-01

    A new type of U-shape anti-cathode X-ray generator in which the inner surface of a cylindrical target is irradiated by an electron beam has been made by modifying a conventional rotating anti-cathode X-ray generator whose brightness in the catalog is 12 kW mm{sup −2}. A brightness of 129 kW mm{sup −2} was thereby obtained with this new U-shape-type X-ray generator. This new X-ray generator is expected to be of keen interest for applications in academia, industry and in hospitals. A new type of U-shape anti-cathode X-ray generator in which the inner surface of a cylindrical target is irradiated by an electron beam has been made by modifying a conventional rotating anti-cathode X-ray generator whose brightness in the catalog is 12 kW mm{sup −2}. The target material (Cu), target radius (50 mm) and rotating speed (6000 r.p.m.) were not changed in this modification. A brightness of 52 kW mm{sup −2} was obtained by this U-shape-type X-ray generator. This means that the brightness of the new type is 4.3 times greater than that of the old unmodified one. Furthermore, the new-type X-ray generator yielded a brightness of 129 kW mm{sup −2} by adding a carbon coating on the Cu target. This means an overall increase of brightness of ten times. The original generator has the highest brightness in the generators of the same class (having a radius of 50 mm and rotation speed of 6000 r.p.m.). Observations showed that Cu Kα counts at vertical incidence of the electron beam onto the surface of the new target, which is initially optically smooth, decrease as the surface is roughened by a severe thermal stress caused by strong electron beam exposure. Further observation reveals, however, that oblique incidence of the electron beam onto the roughened surface drastically increased the X-ray output and amounts to twice as much as that from a smooth surface at vertical incidence. Thus, at the present stage, an overall increase of brightness has been realised at a level 20 times

  5. Compact scanning transmission x-ray microscope at the photon factory

    International Nuclear Information System (INIS)

    Takeichi, Yasuo; Inami, Nobuhito; Ono, Kanta; Suga, Hiroki; Takahashi, Yoshio

    2016-01-01

    We report the design and performance of a compact scanning transmission X-ray microscope developed at the Photon Factory. Piezo-driven linear stages are used as coarse stages of the microscope to realize excellent compactness, mobility, and vibrational and thermal stability. An X-ray beam with an intensity of ∼10 7 photons/s was focused to a diameter of ∼40 nm at the sample. At the soft X-ray undulator beamline used with the microscope, a wide range of photon energies (250–1600 eV) is available. The microscope has been used to research energy materials and in environmental sciences

  6. Picosecond x-ray measurements from 100 eV to 30 keV

    International Nuclear Information System (INIS)

    Attwood, D.T.; Kauffman, R.L.; Stradling, G.L.

    1980-01-01

    Picosecond x-ray measurements relevant to the Livermore Laser Fusion Program are reviewed. Resolved to 15 picoseconds, streak camera detection capabilities extend from 100 eV to higher than 30 keV, with synchronous capabilities in the visible, near infrared, and ultraviolet. Capabilities include automated data retrieval using charge coupled devices (CCD's), absolute x-ray intensity levels, novel cathodes, x-ray mirror/reflector combinations, and a variety of x-ray imaging devices

  7. X-ray backscatter radiography. Intrusive instead of penetrating, X-ray shadow phenomenon

    International Nuclear Information System (INIS)

    Wrobel, Norma; Kolkoori, Sanjeevareddy; Osterloh, Kurt; European Federation for Non-Destructive Testing

    2013-01-01

    Generally, the primary practical advantage of X-ray backscattering radiography is that there is no need to place a detector on the side of the specimen opposite to the source. Such a situation usually is encountered whenever the specimen is not only standing right in front of a wall or even inside a wall but also if the specimen is such big that radiography is not possible because of the layer thickness to be penetrated. The method used here differs fundamentally from the conventional method to interrogate the object with a scanning beam ('pencil beam') and to collect the whole backscattered radiation from the area. The object is fully illuminated by a (uncollimated) cone beam. Here, the image is recorded with a camera of absorbent material (tungsten, lead), which contains a matrix detector as the image receiver. The optical effect is generated by a special twisted slit collimator which operates according to an extended pinhole camera. The independent positioning of source and camera allows a variable irradiation geometry which causes different images as a result. As a consequence, a complex object in front of a backscattering wall appears completely different than standing alone. So X-ray backscatter images have to be interpreted according to their illumination with X-rays and their surroundings. (orig.)

  8. Modelling properties of hard x-rays generated by the interaction between relativistic electrons and very intense laser beams

    International Nuclear Information System (INIS)

    Popa, Alexandru

    2009-01-01

    In a previous paper we presented a calculation model for high harmonic generation by relativistic Thomson scattering of the electromagnetic radiation by free electrons. In this paper we present a similar model for the calculation of the energies of hard x-rays (20- 200 keV) resulted from the interaction between relativistic electrons (20-100 MeV) and very intense laser beams. Starting from the relativistic equations of motion of an electron in the electromagnetic field we show that the Lienard-Wiechert equation leads to electromagnetic waves whose frequencies are in the domain of hard x-rays. When the relativistic parameter of the laser beam is greater than unity, the model predicts the existence of harmonics of the above frequencies. Our theoretical values are in good agreement with experimental values of the x-ray energies from the literature and predict accurately their angular distribution.

  9. Development and characterization of a tunable ultrafast X-ray source via inverse-Compton-scattering

    International Nuclear Information System (INIS)

    Jochmann, Axel

    2014-01-01

    Ultrashort, nearly monochromatic hard X-ray pulses enrich the understanding of the dynamics and function of matter, e.g., the motion of atomic structures associated with ultrafast phase transitions, structural dynamics and (bio)chemical reactions. Inverse Compton backscattering of intense laser pulses from relativistic electrons not only allows for the generation of bright X-ray pulses which can be used in a pump-probe experiment, but also for the investigation of the electron beam dynamics at the interaction point. The focus of this PhD work lies on the detailed understanding of the kinematics during the interaction of the relativistic electron bunch and the laser pulse in order to quantify the influence of various experiment parameters on the emitted X-ray radiation. The experiment was conducted at the ELBE center for high power radiation sources using the ELBE superconducting linear accelerator and the DRACO Ti:sapphire laser system. The combination of both these state-of-the-art apparatuses guaranteed the control and stability of the interacting beam parameters throughout the measurement. The emitted X-ray spectra were detected with a pixelated detector of 1024 by 256 elements (each 26μm by 26μm) to achieve an unprecedented spatial and energy resolution for a full characterization of the emitted spectrum to reveal parameter influences and correlations of both interacting beams. In this work the influence of the electron beam energy, electron beam emittance, the laser bandwidth and the energy-anglecorrelation on the spectra of the backscattered X-rays is quantified. A rigorous statistical analysis comparing experimental data to ab-initio 3D simulations enabled, e.g., the extraction of the angular distribution of electrons with 1.5% accuracy and, in total, provides predictive capability for the future high brightness hard X-ray source PHOENIX (Photon electron collider for Narrow bandwidth Intense X-rays) and potential all optical gamma-ray sources. The results

  10. X-ray examination equipment for heart diagnostics

    International Nuclear Information System (INIS)

    Kok, P.W.

    1979-01-01

    For heart catheterization the X-ray tube and the image intensifier can be shifted parallel to the scanning plane. Without moving the patient it is also possible to displace the system X-ray tube/image intensifier arbitrarily in space, while keeping up the direction of the X-ray beam. (RW) [de

  11. A simulation-based study on the influence of beam hardening in X-ray computed tomography for dimensional metrology.

    Science.gov (United States)

    Lifton, Joseph J; Malcolm, Andrew A; McBride, John W

    2015-01-01

    X-ray computed tomography (CT) is a radiographic scanning technique for visualising cross-sectional images of an object non-destructively. From these cross-sectional images it is possible to evaluate internal dimensional features of a workpiece which may otherwise be inaccessible to tactile and optical instruments. Beam hardening is a physical process that degrades the quality of CT images and has previously been suggested to influence dimensional measurements. Using a validated simulation tool, the influence of spectrum pre-filtration and beam hardening correction are evaluated for internal and external dimensional measurements. Beam hardening is shown to influence internal and external dimensions in opposition, and to have a greater influence on outer dimensions compared to inner dimensions. The results suggest the combination of spectrum pre-filtration and a local gradient-based surface determination method are able to greatly reduce the influence of beam hardening in X-ray CT for dimensional metrology.

  12. Beam Hardening Artifacts: Comparison between Two Cone Beam Computed Tomography Scanners

    Directory of Open Access Journals (Sweden)

    Farzad Esmaeili

    2012-04-01

    Full Text Available Background and aims. At present, cone beam computed tomography (CBCT has become a substitute for computed tomography (CT in dental procedures. The metallic materials used in dentistry can produce artifacts due to the beam hardening phenomenon. These artifacts decrease the quality of images. In the present study, the number of artifacts as a result of beam hardening in the images of dental implants was compared between two NewTom VG and Planmeca Promax 3D Max CBCT machines. Materials and methods. An implant drilling model was used in the present study. The implants (Dentis were placed in the canine, premolar and molar areas. Scanning procedures were carried out by two CBCT machines. The corresponding sections (coronal and axial of the implants were evaluated by two radiologists. The number of artifacts in each image was determined using the scale provided. Mann-Whitney U test was used for two-by-two comparisons at a significance level of P<0.05. Results. There were statistically significant differences in beam hardening artifacts in axial and coronal sections between the two x-ray machines (P<0.001, with a higher quality in the images produced by the NewTom VG. Conclusion. Given the higher quality of the images produced by the NewTom VG x-ray machine, it is recommended for imaging of patients with extensive restorations, multiple prostheses or previous implant treatments.

  13. Beam-beam interaction in e+-e- storage rings

    International Nuclear Information System (INIS)

    Le Duff, J.

    1977-01-01

    Colliding beams in electron-positron storage rings are discussed with particular reference to the space charge forces occuring during beam-beam interactions and their effect on beam current and consequently machine performance (maximum luminosity). The first section deals with linear beam-beam effects and discussses linear tune shift; the second section considers non-linear beam-beam effects and the creation on non-linear resonances. The last section poses questions of the possibility of extrapolating present results to future machines and discusses optimization of storage ring performance. (B.D.)

  14. Exploring vacuum birefringence based on a 100 PW laser and an x-ray free electron laser beam

    Science.gov (United States)

    Shen, Baifei; Bu, Zhigang; Xu, Jiancai; Xu, Tongjun; Ji, Liangliang; Li, Ruxin; Xu, Zhizhan

    2018-04-01

    Exploring vacuum birefringence with the station of extreme light at Shanghai Coherent Light Facility is considered. Laser pulses of intensity beyond 1023 W cm-2 are capable of polarizing the vacuum due to the ultra-strong electro-magnetic fields. The subtle difference of the vacuum refractive indexes along electric and magnetic fields leads to a birefringence effect for lights propagating through. The vacuum birefringence effect can now be captured by colliding a hard x-ray free electron laser (XFEL) beam with a high-power laser. The initial XFEL beam of pure linear polarization is predicated to gain a very small ellipticity after passing through the laser stimulated vacuum. Various interaction geometries are considered, showing that the estimated ellipticity lies between 1.8 × 10-10 and 10-9 for a 100 PW laser interacting with a 12.9 keV XFEL beam, approaching the threshold for todays’ polarity detection technique. The detailed experimental set-up is designed, including the polarimeter, the focusing compound refractive lens and the optical path. When taking into account the efficiencies of the x-ray instruments, it is found that about 10 polarization-flipped x-ray photons can be detected for a single shot for our design. Considering the background noise level, accumulating runs are necessary to obtain high confident measurement.

  15. Development, Beam characterization and chromosomal effectiveness of X-rays of RBC characteristic X-ray generator

    International Nuclear Information System (INIS)

    Endo, Satoru; Hoshi, Masaharu; Takada, Jun; Takatsuji, Toshihiro; Ejima, Yosuke; Saigusa, Shin; Tachibana, Akira; Sasaki, Masao S.

    2006-01-01

    A characteristic hot-filament type X-ray generator was constructed for irradiation of cultured cells. The source provides copper K, iron K, chromium K, molybdenum L, aluminium K and carbon K shell characteristic X-rays. When cultured mouse m5S cells were irradiated and frequencies of dicentrics were fitted to a linear-quadratic model, Y=αD+βD 2 , the chromosomal effectiveness was not a simple function of photon energy. The α-terms increased with the decrease of the photon energy and then decreased with further decrease of the energy with an inflection point at around 10 keV. The β-terms stayed constant for the photon energy down to 10 keV and then increased with further decrease of energy. Below 10 keV, the relative biological effectiveness (RBE) at low doses was proportional to the photon energy, which contrasted to that for high energy X- or γ-rays where the RBE was inversely related with the photon energy. The reversion of the energy dependency occurred at around 1-2 Gy, where the RBE of soft X-rays was insensitive to X-ray energy. The reversion of energy-RBE relation at a moderate dose may shed light on the controversy on energy dependency of RBE of ultrasoft X-rays in cell survival experiments. (author)

  16. A free-electron laser fourth-generation X-ray source

    International Nuclear Information System (INIS)

    Moncton, D. E.

    1999-01-01

    The field of synchrotrons radiation research has grown rapidly over the last 25 years due to both the push of the accelerator and magnet technology that produces the x-ray beams and the pull of the extraordinary scientific research those beams make possible. Three successive generations of synchrotrons radiation facilities have resulted in beam brilliances 11 to 12 orders of magnitude greater than the standard laboratory x-ray tube. However, greater advances can be easily imagined given the fact that x-ray beams from present-day facilities do not exhibit the coherence or time structure so familiar with the.optical laser. Theoretical work over the last ten years or so has pointed to the possibility of generating hard x-ray beams with laser-like characteristics. The concept is based on self-amplified spontaneous emission in free electron lasers. The use of a superconducting linac could produce a major, cost-effective facility that spans wavelengths from the ultraviolet to the hard x-ray regime, simultaneously servicing large numbers experimenters from a wide range of disciplines. As with each past generation of synchrotron facilities, immense new scientific opportunities from fourth-generation sources

  17. [Diffraction gratings used in x-ray spectroscopy]: Final report

    International Nuclear Information System (INIS)

    Smith, H.I.

    1988-01-01

    This subcontract was initiated in order to facilitate the development at MIT of technologies for fabricating the very fine diffraction grating required in x-ray spectroscopy at Lawrence Livermore Laboratory (LLL). These gratings are generally gold transmission gratings with spatial periods of 200 nm or less. The major focus of our efforts was to develop a means of fabricating gratings of 100 nm period. We explored two approaches: e-beam fabrication of x-ray lithography masks, and achromatic holographic lithography. This work was pursued by Erik Anderson as a major component of his Ph.D. thesis. Erik was successful in both the e-beam and holographic approaches. However, the e-beam method proved to be highly impractical: exposure times of about 115 days would be required to cover an area of 1 cm 2 . The achromatic holography, on the other hand, should be capable of exposing areas well in excess of 1 cm 2 in times under 1 hour. Moreover, 100 nm-period gratings produced by achromatic holography are coherent over their entire area whereas gratings produced by e-beam lithography are coherent only over areas /approximately/100 μm. The remainder of this report consists of portions excerpted from Erik Anderson's thesis. These contain all the details of our work on 100 nm period gratings. 26 refs., 17 figs

  18. The toroidicity-induced Alfven eigenmode structure in DIII-D: Implications of soft x-ray and beam-ion loss data

    International Nuclear Information System (INIS)

    Carolipio, E. M.; Heidbrink, W. W.; Cheng, C. Z.; Chu, M. S.; Fu, G. Y.; Jaun, A.; Spong, D. A.; Turnbull, A. D.; White, R. B.

    2001-01-01

    The internal structure of the toroidicity-induced Alfven eigenmode (TAE) is studied by comparing soft x-ray profile and beam ion loss data taken during TAE activity in the DIII-D tokamak [W. W. Heidbrink , Nucl. Fusion 37, 1411 (1997)] with predictions from theories based on ideal magnetohydrodynamic (MHD), gyrofluid, and gyrokinetic models. The soft x-ray measurements indicate a centrally peaked eigenfunction, a feature which is closest to the gyrokinetic model's prediction. The beam ion losses are simulated using a guiding center code. In the simulations, the TAE eigenfunction calculated using the ideal MHD model acts as a perturbation to the equilibrium field. The predicted beam ion losses are an order of magnitude less than the observed ∼6%--8% losses at the peak experimental amplitude of {delta}B r /B 0 ≅2--5 x 10 -4

  19. Moving-Article X-Ray Imaging System and Method for 3-D Image Generation

    Science.gov (United States)

    Fernandez, Kenneth R. (Inventor)

    2012-01-01

    An x-ray imaging system and method for a moving article are provided for an article moved along a linear direction of travel while the article is exposed to non-overlapping x-ray beams. A plurality of parallel linear sensor arrays are disposed in the x-ray beams after they pass through the article. More specifically, a first half of the plurality are disposed in a first of the x-ray beams while a second half of the plurality are disposed in a second of the x-ray beams. Each of the parallel linear sensor arrays is oriented perpendicular to the linear direction of travel. Each of the parallel linear sensor arrays in the first half is matched to a corresponding one of the parallel linear sensor arrays in the second half in terms of an angular position in the first of the x-ray beams and the second of the x-ray beams, respectively.

  20. Spot size characterization of focused non-Gaussian X-ray laser beams

    Czech Academy of Sciences Publication Activity Database

    Chalupský, Jaromír; Krzywinski, J.; Juha, Libor; Hájková, Věra; Cihelka, Jaroslav; Burian, Tomáš; Vyšín, Luděk; Gaudin, J.; Gleeson, A.; Jurek, M.; Khorsand, A.R.; Klinger, D.; Wabnitz, H.; Sobierajski, R.; Störmer, M.; Tiedtke, K.; Toleikis, S.

    2010-01-01

    Roč. 18, č. 26 (2010), s. 27836-27845 ISSN 1094-4087 R&D Projects: GA AV ČR KAN300100702; GA MŠk LC510; GA MŠk(CZ) LC528; GA ČR GAP208/10/2302; GA MŠk LA08024; GA AV ČR IAAX00100903; GA MŠk(CZ) ME10046 Institutional research plan: CEZ:AV0Z10100523 Keywords : X-ray laser * free-electron laser * beam characterization * ablation Subject RIV: BH - Optics, Masers, Lasers Impact factor: 3.749, year: 2010

  1. Hard X-ray test and evaluation of a prototype 32x32 pixel gallium-arsenide array

    International Nuclear Information System (INIS)

    Erd, C.; Owens, A.; Brammertz, G.; Bavdaz, M.; Peacock, A.; Laemsae, V.; Nenonen, S.; Andersson, H.; Haack, N.

    2002-01-01

    We report X-ray measurements on a prototype 1.1 cm 2 , 32x32 GaAs pixel array with a pixel size of 350x350 μm 2 produced to assess the technological feasibility of making large area, almost Fano-limited arrays, which operate near room temperature. Measurements were carried out on four widely separated pixels both in our laboratories and using monochromatic X-ray pencil beams at the HASYLAB synchrotron research facility in Hamburg, Germany. The pixels were found to be very uniform both in their energy and spatial responses. For example, typical energy resolutions of ∼280 eV at 10.5 keV, rising to ∼560 eV at 60 keV were achieved. The corresponding resolutions measured under full-pixel illumination were found to be the same within statistics, indicating uniform crystallinity and stoichiometry. Likewise, by scanning a 15 keV, 15x15 μm 2 beam across the entire surface of each of the pixels, the gain uniformity across the pixels (and by implication the entire array) was determined to be statistically flat

  2. Higher coherent x-ray laser

    International Nuclear Information System (INIS)

    Hasegawa, Noboru; Nagashima, Keisuke; Kawachi, Tetsuya

    2001-01-01

    X-ray lasers generated by an ultra short pulse laser have advantages such as monochromatic, short pulse duration, small beam divergence, high intensity, and coherence. Spatial coherence is most important for applications, we have investigated the transient collisional excitation (TCE) scheme x-ray laser lasing from Ne-like titanium (31.6 nm), Ne-like silver (13.9 nm) and tin (11.9 nm). However, the spatial coherence was not so good with this scheme. We have been studying to improve the spatial coherence of the x-ray laser and have proposed to use coherent seed light tuned to the x-ray laser wavelength generated from higher harmonics generation (HHG), which is introduced to the x-ray laser medium (Ne-like titanium, Ni-like silver plasmas). We present about the theoretical study of the coupling efficiency HHG light with x-ray laser medium. (author)

  3. Development of Compact Soft X-ray Source Based on Laser Undulator

    CERN Document Server

    Kuroda, Ryunosuke; Minamiguchi, S; Saitô, T; Ueyama, D; Washio, Masakazu

    2004-01-01

    A compact soft X-ray source is required in various research fields such as material and biological science. The laser undulator based on backward Compton scattering has been developed as a compact soft X-ray source for the biological observation at Waseda University. It is performed in a water window region (250eV - 500 eV) using the interaction between 1047 nm Nd:YLF laser and 4 MeV high quality electron beam generated from rf gun system. The range of energy in the water window region has K-shell absorption edges of Oxygen, Carbon and Nitrogen, which mainly constitute of living body. Since the absorption coefficient of water is much smaller than the protein’s coefficient in this range, a dehydration of the specimens is not necessary. As a preliminary experiment, about 300 eV X-ray generation was carried out. As next step, soft X-ray optics with zone plate was proposed for Soft X-ray microscopy. In this conference, we will report details and results of the experiment.

  4. X-ray sky

    International Nuclear Information System (INIS)

    Gruen, M.; Koubsky, P.

    1977-01-01

    The history is described of the discoveries of X-ray sources in the sky. The individual X-ray detectors are described in more detail, i.e., gas counters, scintillation detectors, semiconductor detectors, and the principles of X-ray spectrometry and of radiation collimation aimed at increased resolution are discussed. Currently, over 200 celestial X-ray sources are known. Some were identified as nebulae, in some pulsations were found or the source was identified as a binary star. X-ray bursts of novae were also observed. The X-ray radiation is briefly mentioned of spherical star clusters and of extragalactic X-ray sources. (Oy)

  5. X-ray microscopy using collimated and focussed synchrotron radiation

    International Nuclear Information System (INIS)

    Jones, K.W.; Kwiatek, W.M.; Gordon, B.M.

    1987-01-01

    X-ray microscopy is a field that has developed rapidly in recent years. Two different approaches have been used. Zone plates have been employed to produce focused beams with sizes as low as 0.07 μm for x-ray energies below 1 keV. Images of biological materials and elemental maps for major and minor low Z have been produced using above and below absorption edge differences. At higher energies collimators and focusing mirrors have been used to make small diameter beams for excitation of characteristic K- or L-x rays of all elements in the periodic table. The practicality of a single instrument combining all the features of these two approaches is unclear. The use of high-energy x rays for x-ray microscopy has intrinsic value for characterization of thick samples and determination of trace amounts of most elements. A summary of work done on the X-26 beam line at the National Synchrotron Light Source (NSLS) with collimated and focused x rays with energies above 4 keV is given here. 6 refs., 5 figs., 1 tab

  6. Intercomparison of ionization chambers in standard X-ray beams, at radiotherapy, diagnostic radiology and radioprotection levels; Intercomparacao de camaras de ionizacao em feixes padroes de raios X, niveis radioterapia, radiodiagnostico e radioprotecao

    Energy Technology Data Exchange (ETDEWEB)

    Bessa, Ana Carolina Moreira de

    2006-07-01

    Since the calibration of radiation measurement instruments and the knowledge of their major characteristics are very important subjects, several different types of ionization chambers were intercompared in terms of their calibration coefficients and their energy dependence, in radiotherapy, diagnostic radiology and radioprotection standard beams. An intercomparison of radionuclide calibrators for nuclear medicine was performed, using three radionuclides: {sup 67}Ga, {sup 201}Tl and {sup 99m}Tc; the results obtained were all within the requirements of the national standard CNEN-NE-3.05. In order to complete the range of radiation qualities of the Calibration Laboratory of IPEN, standard radiation beam qualities, radiation protection and low energy radiation therapy levels, were established, according international recommendations. Three methodologies for the calibration of unsealed ionization chambers in X-ray beams were studied and compared. A set of Victoreen ionization chambers, specially designed for use in laboratorial intercomparisons, was submitted to characterization tests. The performance of these Victoreen ionization chambers showed that they are suitable for use in radioprotection beams, because the results obtained agree with international recommendations. However, these Victoreen ionization chambers can be used in radiotherapy and diagnostic radiology beams only with some considerations, since their performance in these beams, especially in relation to the energy dependence and stabilization time tests, did not agree with the international recommendations for dosimeters used in radiotherapy and diagnostic radiology beams. This work presents data on the performance of several types of ionization chambers in different X-ray beams, that may be useful for choosing the appropriate instrument for measurements in ionizing radiation beams. (author)

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... x-ray machine is a compact apparatus that can be taken to the patient in a hospital ... so that any change in a known abnormality can be monitored over time. Follow-up examinations are ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... carefully aimed at the part of the body being examined, an x-ray machine produces a small ... the table in the area of the body being imaged. When necessary, sandbags, pillows or other positioning ...

  9. Technique charts for Kodak EC-L film screen system for portal localization in a 6MV X-ray beam.

    Science.gov (United States)

    Sandilos, P; Antypas, C; Paraskevopoulou, C; Kouvaris, J; Vlachos, L

    2006-01-01

    Port films are used in radiotherapy for visual evaluation of the radiation fields and subsequent quantitative analysis. Common port films suffer from poor image quality compared to the simulator-diagnostic films and is desirable to determine the appropriate exposure required for the best image contrast. The aim of this work is to generate technique charts for the Kodak EC-L film screen system for use in a 6MV x-ray beam. Three homogeneous water phantoms were used to simulate head-neck, thorax and abdomen dimensions of adult human, correspondingly. The film screen system was calibrated in a 6MV x-ray beam and under various irradiation conditions. The film screen system behavior was studied as a function of phantom thickness, field size and air gap between the phantom and the film screen system. In each case the optimum film exposure which produces the maximum image contrast was determined. The generated technique charts for the EC-L film screen system and for a 6 MV x-ray beam are used in our radiotherapy department for daily quality assurance of the radiotherapy procedure.

  10. Effective Energy Determination Of Radiodiagnostic X-Rays

    International Nuclear Information System (INIS)

    Sumarni; Mart, Terry

    2000-01-01

    X-rays have been used for diagnostic radiology to produce image on film that give anatomy information. Effective energy should be known to get benefit exposure. Half value layer (HVL) as shown as monoenergetic x-rays has similar spectra of energy x-rays. It has been done measurement with x-ray machine Tanka at P3KRBIN-Batan for 40 kVp to 119 kVp of potential found of Aluminium HVL are 0.115 cm to 0.385 cm and energy effective between 23.24 keV to 37.5 keV

  11. Crystal nucleation in metallic alloys using x-ray radiography and machine learning

    Science.gov (United States)

    Arteta, Carlos; Lempitsky, Victor

    2018-01-01

    The crystallization of solidifying Al-Cu alloys over a wide range of conditions was studied in situ by synchrotron x-ray radiography, and the data were analyzed using a computer vision algorithm trained using machine learning. The effect of cooling rate and solute concentration on nucleation undercooling, crystal formation rate, and crystal growth rate was measured automatically for thousands of separate crystals, which was impossible to achieve manually. Nucleation undercooling distributions confirmed the efficiency of extrinsic grain refiners and gave support to the widely assumed free growth model of heterogeneous nucleation. We show that crystallization occurred in temporal and spatial bursts associated with a solute-suppressed nucleation zone. PMID:29662954

  12. Multiple station beamline at an undulator x-ray source

    DEFF Research Database (Denmark)

    Als-Nielsen, J.; Freund, A.K.; Grübel, G.

    1994-01-01

    The undulator X-ray source is an ideal source for many applications: the beam is brilliant, highly collimated in all directions, quasi-monochromatic, pulsed and linearly polarized. Such a precious source can feed several independently operated instruments by utilizing a downstream series of X......-ray transparent monochromator crystals. Diamond in particular is an attractive monochromator as it is rather X-ray transparent and can be fabricated to a high degree of crystal perfection. Moreover, it has a very high heat conductivity and a rather small thermal expansion so the beam X-ray heat load problem...

  13. X-ray total reflection mirrors for coherent illumination

    CERN Document Server

    Ishikawa, T; Yabashi, M; Souvorov, A; Yamauchi, K; Yamamura, K; Mimura, H; Saito, A; Mori, Y

    2002-01-01

    X-ray mirrors for coherent illumination demand much higher surface quality than is achievable with the conventional polishing techniques. Plasma chemical vaporization machining (CVM) and elastic emission machining (EEM) have been applied for x-ray mirror manufacturing. Figure error of a flat silicon single crystal mirrors made with CVM+EEM process was reduced to 2.0 nm peak-to-valley and 0.2 nm RMS. The machining process was also applied to make elliptical mirrors. One-dimensional focusing with a single elliptical mirror showed diffraction-limited properties with the focal width of 200 nm. Two-dimensional focusing with Kirkpatric-Baez configuration gave a focal spot size of 200 nm x 200 nm. (author)

  14. Fluorescence-type Monochromatic X-ray Beam-position Monitor with High-spatial Resolution for the NSLS-II Beamlines

    International Nuclear Information System (INIS)

    Yoon, Phil S.; Siddons, D. Peter

    2010-01-01

    We developed a fluorescence-type monochromatic X-ray beam-position monitor (X-BPM) with high-spatial resolution for end-station experiments at the initial project beamlines of the NSLS-II. We designed a ring array of multi-segmented Si PIN-junction photodiodes to use as a position sensor. Further, we integrated a low-noise charge-preamplification HERMES4 ASIC chip into an electronic readout system for photon-counting application. A series of precision measurements to characterize electronically the Si-photodiode sensor and the ASIC chip demonstrated that the inherent noise from the detector system is sufficiently low to meet our stringent requirements. Using a Gaussian beam, we parametrically modeled the optimum working distance to ensure the detector's best performance. Based upon the results from the parametric modeling, prototypes of the next versions of the X-BPM are being developed. In this paper, we describe the methodology for developing the new compact monochromatic X-ray BPM, including its instrumentation, detector modeling, and future plan.

  15. An experimental and Monte Carlo investigation of the energy dependence of alanine/EPR dosimetry: I. Clinical x-ray beams

    International Nuclear Information System (INIS)

    Zeng, G G; McEwen, M R; Rogers, D W O; Klassen, N V

    2004-01-01

    The energy dependence of alanine/EPR dosimetry, in terms of absorbed dose-to-water for clinical 6, 10, 25 MV x-rays and 60 Co rays was investigated by measurements and Monte Carlo (MC) calculations. The dose rates were traceable to the NRC primary standard for absorbed dose, a sealed water calorimetry. The electron paramagnetic resonance (EPR) spectra of irradiated pellets were measured using a Bruker EMX 081 EPR spectrometer. The DOSRZnrc Monte Carlo code of the EGSnrc system was used to simulate the experimental conditions with BEAM code calculated input spectra of x-rays and γ-rays. Within the experimental uncertainty of 0.5%, the alanine EPR response to absorbed dose-to-water for x-rays was not dependent on beam quality from 6 MV to 25 MV, but on average, it was about 0.6% lower than its response to 60 Co gamma rays. Combining experimental data with Monte Carlo calculations, it is found that the alanine/EPR response per unit absorbed dose-to-alanine is the same for clinical x-rays and 60 Co gamma rays within the uncertainty of 0.6%. Monte Carlo simulations showed that neither the presence of PMMA holder nor varying the dosimeter thickness between 1 mm and 5 mm has significant effect on the energy dependence of alanine/EPR dosimetry within the calculation uncertainty of 0.3%

  16. X-ray simulation development

    International Nuclear Information System (INIS)

    Posey, L.D.; Tollefsrud, P.B.; Woodall, H.W.; McDaniel, D.H.; Allred, R.E.

    1975-01-01

    Design modifications are discussed for an electron beam accelerator used as a Bremsstrahlung x-ray source. The primary goal of the program, to obtain a reliable 5 cal/gm exposure capability, can be accomplished with beam compression by an external magnetic guide field. Initial operating characteristics and performance improvements are presented

  17. Refractive optics to compensate x-ray mirror shape-errors

    Science.gov (United States)

    Laundy, David; Sawhney, Kawal; Dhamgaye, Vishal; Pape, Ian

    2017-08-01

    Elliptically profiled mirrors operating at glancing angle are frequently used at X-ray synchrotron sources to focus X-rays into sub-micrometer sized spots. Mirror figure error, defined as the height difference function between the actual mirror surface and the ideal elliptical profile, causes a perturbation of the X-ray wavefront for X- rays reflecting from the mirror. This perturbation, when propagated to the focal plane results in an increase in the size of the focused beam. At Diamond Light Source we are developing refractive optics that can be used to locally cancel out the wavefront distortion caused by figure error from nano-focusing elliptical mirrors. These optics could be used to correct existing optical components on synchrotron radiation beamlines in order to give focused X-ray beam sizes approaching the theoretical diffraction limit. We present our latest results showing measurement of the X-ray wavefront error after reflection from X-ray mirrors and the translation of the measured wavefront into a design for refractive optical elements for correction of the X-ray wavefront. We show measurement of the focused beam with and without the corrective optics inserted showing reduction in the size of the focus resulting from the correction to the wavefront.

  18. Nanoscale radiation transport and clinical beam modeling for gold nanoparticle dose enhanced radiotherapy (GNPT) using X-rays.

    Science.gov (United States)

    Zygmanski, Piotr; Sajo, Erno

    2016-01-01

    We review radiation transport and clinical beam modelling for gold nanoparticle dose-enhanced radiotherapy using X-rays. We focus on the nanoscale radiation transport and its relation to macroscopic dosimetry for monoenergetic and clinical beams. Among other aspects, we discuss Monte Carlo and deterministic methods and their applications to predicting dose enhancement using various metrics.

  19. Gamma-H2AX foci in cells exposed to a mixed beam of X-rays and alpha particles

    Science.gov (United States)

    2012-01-01

    Background Little is known about the cellular effects of exposure to mixed beams of high and low linear energy transfer radiation. So far, the effects of combined exposures have mainly been assessed with clonogenic survival or cytogenetic methods, and the results are contradictory. The gamma-H2AX assay has up to now not been applied in this context, and it is a promising tool for investigating the early cellular response to mixed beam irradiation. Purpose To determine the dose response and repair kinetics of gamma-H2AX ionizing radiation-induced foci in VH10 human fibroblasts exposed to mixed beams of 241Am alpha particles and X-rays. Results VH10 human fibroblasts were irradiated with each radiation type individually or both in combination at 37°C. Foci were scored for repair kinetics 0.5, 1, 3 and 24 h after irradiation (one dose per irradiation type), and for dose response at the 1 h time point. The dose response effect of mixed beam was additive, and the relative biological effectiveness for alpha particles (as compared to X-rays) was of 0.76 ± 0.52 for the total number of foci, and 2.54 ± 1.11 for large foci. The repair kinetics for total number of foci in cells exposed to mixed beam irradiation was intermediate to that of cells exposed to alpha particles and X-rays. However, for mixed beam-irradiated cells the frequency and area of large foci were initially lower than predicted and increased during the first 3 hours of repair (while the predicted number and area did not). Conclusions The repair kinetics of large foci after mixed beam exposure was significantly different from predicted based on the effect of the single dose components. The formation of large foci was delayed and they did not reach their maximum area until 1 h after irradiation. We hypothesize that the presence of low X-ray-induced damage engages the DNA repair machinery leading to a delayed DNA damage response to the more complex DNA damage induced by alpha particles. PMID:23121736

  20. Intercomparison of ionization chambers in standard X-ray beams, at radiotherapy, diagnostic radiology and radioprotection levels

    International Nuclear Information System (INIS)

    Bessa, Ana Carolina Moreira de

    2007-01-01

    Since the calibration of radiation measurement instruments and the knowledge of their major characteristics are very important subjects, several different types of ionization chambers were inter compared in terms of their calibration coefficients and their energy dependence, in radiotherapy, diagnostic radiology and radioprotection standard beams. An intercomparison of radionuclide calibrators for nuclear medicine was performed, using three radionuclides: 67 Ga, 201 Tl and 99m Tc; the results obtained were all within the requirements of the national standard CNEN-NE-3.05. In order to complete the range of radiation qualities of the Calibration Laboratory of IPEN, standard radiation beam qualities, radiation protection and low energy radiation therapy levels, were established, according international recommendations. Three methodologies for the calibration of unsealed ionization chambers in X-ray beams were studied and compared. A set of Victoreen ionization chambers, specially designed for use in laboratorial intercomparisons, was submitted to characterization tests. The performance of these Victoreen ionization chambers showed that they are suitable for use in radioprotection beams, because the results obtained agree with international recommendations. However, these Victoreen ionization chambers can be used in radiotherapy and diagnostic radiology beams only with some considerations, since their performance in these beams, especially in relation to the energy dependence and stabilization time tests, did not agree with the international recommendations for dosimeters used in radiotherapy and diagnostic radiology beams. This work presents data on the performance of several types of ionization chambers in different X-ray beams, that may be useful for choosing the appropriate instrument for measurements in ionizing radiation beams. (author)