  1. Independent technique of verifying high-dose rate (HDR) brachytherapy treatment plans

    International Nuclear Information System (INIS)

    Saw, Cheng B.; Korb, Leroy J.; Darnell, Brenda; Krishna, K. V.; Ulewicz, Dennis

    1998-01-01

    Purpose: An independent technique for verifying high-dose rate (HDR) brachytherapy treatment plans has been formulated and validated clinically. Methods and Materials: In HDR brachytherapy, dwell times at respective dwell positions are computed, using an optimization algorithm in a HDR treatment-planning system to deliver a specified dose to many target points simultaneously. Because of the variability of dwell times, concerns have been expressed regarding the ability of the algorithm to compute the correct dose. To address this concern, a commercially available low-dose rate (LDR) algorithm was used to compute the doses at defined distances, based on the dwell times obtained from the HDR treatment plans. The percent deviation between doses computed using the HDR and LDR algorithms were reviewed for HDR procedures performed over the last year. Results: In this retrospective study, the difference between computed doses using the HDR and LDR algorithms was found to be within 5% for about 80% of the HDR procedures. All of the reviewed procedures have dose differences of less than 10%. Conclusion: An independent technique for verifying HDR brachytherapy treatment plans has been validated based on clinical data. Provided both systems are available, this technique is universal in its applications and not limited to either a particular implant applicator, implant site, or implant type

  2. Dose rate considerations in brachytherapy: biological equivalence of LDR and HDR

    International Nuclear Information System (INIS)

    Orton, C.G.

    1994-01-01

    The linear-quadratic model for cell survival and bioeffect doses is discussed and equations for low dose rate (LDR), high dose rate (HDR) and intermediate situations are presented. The model, when used to define LDR and single fractions of HDR, shows, that these correspond to irradiations lasting longer than about 14 hours or shorter than about 0.7 hours, respectively. It is shown that, for HDR to be as safe and effective as LDR, the dose-rate effect of LDR has to be replaced by the fractionation-effect of HDR. This is necessary in order to take advantage of the differential repair characteristics between late-reacting normal tissue and tumor cells at low doses and low dose rates. Using the linear-quadratic model to simulate repair mathematically, it is shown that the number of fractions required is highly dependent upon what parameters are assumed for normal tissues and tumor, as well as whether or not there is any physical advantage gained by conversion from LDR to HDR. (author). 20 refs., 7 figs

  3. Developing A Directional High-Dose Rate (d-HDR) Brachytherapy Source

    Science.gov (United States)

    Heredia, Athena Yvonne

    Conventional sources used in brachytherapy provide nearly isotropic or radially symmetric dose distributions. Optimizations of dose distributions have been limited to varied dwell times at specified locations within a given treatment volume, or manipulations in source position for seed implantation techniques. In years past, intensity modulated brachytherapy (IMBT) has been used to reduce the amount of radiation to surrounding sensitive structures in select intracavitary cases by adding space or partial shields. Previous work done by Lin et al., at the University of Wisconsin-Madison, has shown potential improvements in conformality for brachytherapy treatments using a directionally shielded low dose rate (LDR) source for treatments in breast and prostate. Directional brachytherapy sources irradiate approximately half of the radial angles around the source, and adequately shield a quarter of the radial angles on the opposite side, with sharp gradient zones between the treated half and shielded quarter. With internally shielded sources, the radiation can be preferentially emitted in such a way as to reduce toxicities in surrounding critical organs. The objective of this work is to present findings obtained in the development of a new directional high dose rate (d-HDR) source. To this goal, 103Pd (Z = 46) is reintroduced as a potential radionuclide for use in HDR brachytherapy. 103Pd has a low average photon energy (21 keV) and relatively short half -life (17 days), which is why it has historically been used in low dose rate applications and implantation techniques. Pd-103 has a carrier-free specific activity of 75000 Ci/g. Using cyclotron produced 103Pd, near carrier-free specific activities can be achieved, providing suitability for high dose rate applications. The evolution of the d-HDR source using Monte Carlo simulations is presented, along with dosimetric parameters used to fully characterize the source. In addition, a discussion on how to obtain elemental

  4. High dose rate (HDR) and low dose rate (LDR) interstitial irradiation (IRT) of the rat spinal cord

    International Nuclear Information System (INIS)

    Pop, Lucas A.M.; Plas, Mirjam van der; Skwarchuk, Mark W.; Hanssen, Alex E.J.; Kogel, Albert J. van der

    1997-01-01

    Purpose: To describe a newly developed technique to study radiation tolerance of rat spinal cord to continuous interstitial irradiation (IRT) at different dose rates. Material and methods: Two parallel catheters are inserted just laterally on each side of the vertebral bodies from the level of Th 10 to L 4 . These catheters are afterloaded with two 192 Ir wires of 4 cm length each (activity 1-2.3 mCi/cm) for the low dose rate (LDR) IRT or connected to the HDR micro-Selectron for the high dose rate (HDR) IRT. Spinal cord target volume is located at the level of Th 12 -L 2 . Due to the rapid dose fall-off around the implanted sources, a dose inhomogeneity across the spinal cord thickness is obtained in the dorso-ventral direction. Using the 100% reference dose (rate) at the ventral side of the spinal cord to prescribe the dose, experiments have been carried out to obtain complete dose response curves at average dose rates of 0.49, 0.96 and 120 Gy/h. Paralysis of the hind-legs after 5-6 months and histopathological examination of the spinal cord of each irradiated rat are used as experimental endpoints. Results: The histopathological damage seen after irradiation is clearly reflected the inhomogeneous dose distribution around the implanted catheters, with the damage predominantly located in the dorsal tract of the cord or dorsal roots. With each reduction in average dose rate, spinal cord radiation tolerance is significantly increased. When the dose is prescribed at the 100% reference dose rate, the ED 50 (induction of paresis in 50% of the animals) for the HDR-IRT is 17.3 Gy. If the average dose rate is reduced from 120 Gy/h to 0.96 or 0.49 Gy/h, a 2.9- or 4.7-fold increase in the ED 50 values to 50.3 Gy and 80.9 Gy is observed; for the dose prescribed at the 150% reference dose rate (dorsal side of cord) ED 50 values are 26.0, 75.5 and 121.4 Gy, respectively. Using different types of analysis and in dependence of the dose prescription and reference dose rate, the Î

  5. Dose rate effect from the relationship between ICRU rectal dose and local control rate in intracavitary radiotherapy for carcinoma of the uterine cervix. Six fraction HDR and three-fraction LDR in three weeks

    International Nuclear Information System (INIS)

    Jingu, Kenichi; Akita, Yuzou; Ohmagari, Jyunichi

    2001-01-01

    The dose rate effect, low dose rate radiotherapy (LDR)/high dose rate radiotherapy (HDR), was calculated using the isoeffect ICRU rectal dose by intracavitary radiotherapy (ICRT) for uterine cervix cancer. The subjects analyzed consisted of 78 LDR and 74 HDR patients whose ICRU rectal dose could be calculated and whose local control as stage II/III cases could be evaluated. The point A dose in ICRT was 45-55 Gy/3 fractions/3 weeks for LDR and 30 Gy/6 fractions/3 weeks for HDR. The dose effect relationships associated with local control at each whole pelvis external radiation dose were calculated using the double integration method and Probit analysis, and the 50% and 90% local control ICRU rectal doses were calculated from this relationship. Finally, the dose rate effect LDR/HDR was determined from 50% and 90% local control doses. The dose rate effect calculated from the 50% local control dose was 1.24 and that from the 90% local control dose was 1.14. (author)

  6. Independent verification of the delivered dose in High-Dose Rate (HDR) brachytherapy

    International Nuclear Information System (INIS)

    Portillo, P.; Feld, D.; Kessler, J.

    2009-01-01

    An important aspect of a Quality Assurance program in Clinical Dosimetry is an independent verification of the dosimetric calculation done by the Treatment Planning System for each radiation treatment. The present paper is aimed at creating a spreadsheet for the verification of the dose recorded at a point of an implant with radioactive sources and HDR in gynecological injuries. An 192 Ir source automatic differed loading equipment, GammaMedplus model, Varian Medical System with HDR installed at the Angel H. Roffo Oncology Institute has been used. The planning system implemented for getting the dose distribution is the BraquiVision. The sources coordinates as well as those of the calculation point (Rectum) are entered into the Excel-devised verification program by assuming the existence of a point source in each one of the applicators' positions. Such calculation point has been selected as the rectum is an organ at risk, therefore determining the treatment planning. The dose verification is performed at points standing at a sources distance having at least twice the active length of such sources, so they may be regarded as point sources. Most of the sources used in HDR brachytherapy with 192 Ir have a 5 mm active length for all equipment brands. Consequently, the dose verification distance must be at least of 10 mm. (author)

  7. Brachytherapy for Buccal Cancer: From Conventional Low Dose Rate (LDR) or Mold Technique to High Dose Rate Interstitial Brachytherapy (HDR-ISBT).

    Science.gov (United States)

    Kotsuma, Tadayuki; Yamazaki, Hideya; Masui, Koji; Yoshida, Ken; Shimizutani, Kimishige; Akiyama, Hironori; Murakami, Shumei; Isohashi, Fumiaki; Yoshioka, Yasuo; Ogawa, Kazuhiko; Tanaka, Eiichi

    2017-12-01

    To examine the effectiveness of newly-installed high-dose-rate interstitial brachytherapy (HDR-ISBT) for buccal cancer. We retrospectively reviewed 36 patients (25 men and 11 women) with buccal cancer treated with curative brachytherapy with or without external radiotherapy with a median follow-up of 99 months. A total of 15 HDR-ISBT (median 48 Gy/ 8 fractions, range=24-60 Gy) patients were compared to conventional 15 cases LDR-ISBT (70 Gy, range=42.8-110 Gy) and 7 molds techniques (15 Gy, range=9-74 Gy). A total of 31 patients also underwent external radiotherapy (30 Gy, range=24-48 Gy). They comprised of 3T1, 23 T2, 8 T3, 3 T4 including 11 node positive cases. HDR-ISBT provided 82% of local control rate at 5 years, whereas conventional brachytherapy showed 72% [p=0.44; LDR-ISBT (65%), mold therapy (85.7%)]. Patients with early lesions (T1-2 or stage I-II) showed better local control rates than those with advanced lesions (T3-4 or stage III-IV). Severe late grade 3 complications developed in two patients treated with LDR-ISBT and EBRT. There is no significant difference in toxicity grade ≤2 between conventional brachytherapy (5/15=33%) and HDR-ISBT (7/32=32%, p=0.92). HDR-ISBT achieved good and comparable local control rates to conventional brachytherapy without elevating the toxicity. Copyright© 2017, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved.

  8. Implementation of microsource high dose rate (mHDR) brachytherapy in developing countries

    International Nuclear Information System (INIS)

    2001-11-01

    Brachytherapy using remote afterloading of a single high dose rate 192 Ir microsource was developed in the 1970s. After its introduction to clinics, this system has spread rapidly among developed Member States and has become a highly desirable modality in cancer treatment. This technique is now gradually being introduced to the developing Member States. The 192 Ir sources are produced with a high specific activity. This results in a high dose rate (HDR) to the tumour and shorter treatment times. The high specific activity simultaneously results in a much smaller source (so-called micro source, around I mm in diameter) which may be easily inserted into tissue through a thin delivery tube, the so-called interstitial treatment, as well as easily inserted into body cavities, the so-called intracavitary or endoluminal treatment. Another advantage is the ability to change dwell time (the time a source remains in one position) of the stepping source which allows dose distribution to match the target volume more closely. The purpose of this TECDOC is to advise radiation oncologists, medical physicists and hospital administrators in hospitals which are planning to introduce 192 Ir microsource HDR (mHDR) remote afterloading systems. The document supplements IAEA-TECDOC-1040, Design and Implementation of a Radiotherapy Programme: Clinical, Medical Physics, Radiation Protection and Safety Aspects, and will facilitate implementation of this new brachytherapy technology, especially in developing countries. The operation of the system, 'how to use the system', is not within the scope of this document. This TECDOC is based on the recommendations of an Advisory Group meeting held in Vienna in April 1999

  9. Comparison of high dose rate (HDR) and low dose rate (LDR) brachytherapy in the treatment of stage IIIB cervix cancer with radiation therapy alone. The preliminary results

    International Nuclear Information System (INIS)

    Trippe, Nivaldo; Novaes, P.E.; Ferrigno, R.; Pellizzon, A.C.; Salvajoli, J.V.; Fogaroli, R.C.; Maia, M.A.C.; Baraldi, H.E.

    1996-01-01

    Purpose/Objective: To compare the results between HDR and LDR brachytherapy in the treatment of stage IIIB cervix cancer with radiation therapy alone through a prospective and randomized trial. Materials and Methods: From September 1992 to December 1993, 65 patients with stage IIIB cervical cancer were randomized to one of the following treatment schedule according to the brachytherapy used to complement the dose of external beam radiotherapy (EBRT): 1 - High dose rate (HDR) - 36 patients - 4 weekly insertions of 6,0 Gy at point A 2 - Low dose rate (LDR) - 29 patients - 2 insertions two weeks apart of 17,5 Gy at point A The External Beam radiotherapy was performed through a Linac 4MV, in box arrangement for whole pelvis and in AP-PA fields for parametrial complementation of dose. The dose at the whole pelvis was 45 Gy in 25 fractions of 1,8 Gy and the parametrial dose was 16 Gy. The brachytherapy was realized with Fletcher colpostats and intrauterine tandem, in both arms. The HDR brachytherapy was realized through a Micro-Selectron device, working with Iridium-192 with initial activity of 10 Ci and started ten days after the beginning of EBRT. The total treatment time was shortened in two weeks for this group. The LDR brachytherapy started only after the end of EBRT. Results: With the minimum follow up of 24 months and medium of 31 months, the disease free survival was 50% among the 36 patients in HDR group and 47,8% among the 29 patients in LDR group. Local failures occurred in 50% and 52,8% respectively. Grade I and II complications were restricted to rectites and cistites and the incidence of them was 8,3% for HDR group and 13% for LDR group. Until the time of evaluation there were no grade III complications in any group. Conclusions: Although the number of patients is small and the time of follow up still short, these preliminary results suggest that the HDR brachytherapy has an equivalent efficiency in local control as the LDR in the treatment of stage IIIB

  10. Perioperative high dose rate (HDR brachytherapy in unresectable locally advanced pancreatic tumors

    Directory of Open Access Journals (Sweden)

    Brygida Białas

    2011-07-01

    Full Text Available Purpose: The aim of the study was to present an original technique of catheter implantation for perioperative HDR-Ir192 brachytherapy in patients after palliative operations of unresectable locally advanced pancreatic tumors and to estimate the influence of perioperative HDR-Ir192 brachytherapy on pain relief in terminal pancreatic cancer patients. Material and methods: Eight patients with pancreatic tumors located in the head of pancreas underwent palliative operations with the use of HDR-Ir192 brachytherapy. All patients qualified for surgery reported pain of high intensity and had received narcotic painkillers prior to operation. During the last phase of the surgery, the Nucletron® catheters were implanted in patients to prepare them for later perioperative brachytherapy. Since the 6th day after surgery HDR brachytherapy was performed. Before each brachytherapy fraction the location of implants were checked using fluoroscopy. A fractional dose was 5 Gy and a total dose was 20 Gy in the area of radiation. A comparative study of two groups of patients (with and without brachytherapy with stage III pancreatic cancer according to the TNM scale was taken in consideration. Results and Conclusions: The authors claim that the modification of catheter implantation using specially designed cannula, facilitates the process of inserting the catheter into the tumor, shortens the time needed for the procedure, and reduces the risk of complications. Mean survival time was 5.7 months. In the group of performed brachytherapy, the mean survival time was 6.7 months, while in the group of no brachytherapy performed – 4.4 months. In the group of brachytherapy, only one patient increased the dose of painkillers in the last month of his life. Remaining patients took constant doses of medicines. Perioperative HDR-Ir192 brachytherapy could be considered as a practical application of adjuvant therapy for pain relief in patients with an advanced pancreatic cancer.

  11. Brachytherapy for early oral tongue cancer. Low dose rate to high dose rate

    International Nuclear Information System (INIS)

    Yamazaki, Hideya; Inoue, Takehiro; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Inoue, Toshihiko; Furukawa, Souhei; Kakimoto, Naoya

    2003-01-01

    To examine the compatibility of low dose rate (LDR) with high dose rate (HDR) brachytherapy, we reviewed 399 patients with early oral tongue cancer (T1-2N0M0) treated solely by brachytherapy at Osaka University Hospital between 1967 and 1999. For patients in the LDR group (n=341), the treatment sources consisted of Ir-192 pin for 227 patients (1973-1996; irradiated dose, 61-85 Gy; median, 70 Gy), Ra-226 needle for 113 patients (1967-1986; 55-93 Gy; median, 70 Gy). Ra-226 and Ir-192 were combined for one patient. Ir-192 HDR (microSelectron-HDR) was used for 58 patients in the HDR group (1991-present; 48-60 Gy; median, 60 Gy). LDR implantations were performed via oral and HDR via a submental/submandibular approach. The dose rates at the reference point for the LDR group were 0.30 to 0.8 Gy/h, and for the HDR group 1.0 to 3.4 Gy/min. The patients in the HDR group received a total dose of 48-60 Gy (8-10 fractions) during one week. Two fractions were administered per day (at least a 6-h interval). The 3- and 5-year local control rates for patients in the LDR group were 85% and 80%, respectively, and those in the HDR group were both 84%. HDR brachytherapy showed the same lymph-node control rate as did LDR brachytherapy (67% at 5 years). HDR brachytherapy achieved the same locoregional result as did LDR brachytherapy. A converting factor of 0.86 is applicable for HDR in the treatment of early oral tongue cancer. (author)

  12. Quality assurance of Vari-source high dose rate (HDR) brachytherapy- remote after loader and cost effectiveness of Vari-source HDR- brachytherapy: NORI, Islamabad experience

    International Nuclear Information System (INIS)

    Ahmad, N.; Mahmood, H.; Jafri, S.R.A.

    2004-01-01

    A quality control of Vari-Source high dose rate (HDR) remote after loading brachytherapy machine was carried out and the cost effectiveness of HDR brachytherapy machine was also evaluated considering the cost of ten Iridium-192 wire sources at Nuclear Medicine, Oncology and Radiotherapy Institute (NORI), Islamabad, Pakistan. A total number of 253 intracavitary insertions were done in 98 patients from October 1996 to May 2001. The results of the quality control tests performed during 1996 to 2001 were within the acceptable limits. The cost effectiveness of Vari-Source HDR brachytherapy machine was also evaluated. The average cost per patient was calculated as US$ 491. Small number of patients was treated as the machine was used for gynecologic malignancies only. The objective was to assess the quality control status of HDR brachytherapy machine on patient treatment day, source exchange day and periodic day (monthly basis). It was found that the cost per patient can be minimized if other type of cancer patients are also treated on Vari-Source HDR machine. (author)

  13. Low dose rate and high dose rate intracavitary treatment for cervical cancer

    International Nuclear Information System (INIS)

    Hareyama, Masato; Oouchi, Atsushi; Shidou, Mitsuo

    1997-01-01

    From 1984 through 1993, 144 previous untreated patients with carcinoma of uterine cervix were treated with either low dose rate 137 Cs therapy (LDR) or high dose rate 60 Co therapy (HDR). The local failure rates for more than 2-years for the primary lesions were 11.8% (8 of 63 patients) for LDR and 18.0% (11 of 61 patients). Rectal complication rates were significantly lower for HDR versus LDR (14.3% VS. 32.8%. p<0.01). Also, bladder complication rates were significantly lower for HDR versus LDR (0% VS. 10.4%, p<0.005). Treatment results in term of local control were equivalent for HDR and LDR treatment. However, the incidence of complications was higher for the LDR group than for the HDR group. (author)

  14. Mathematical simulation of biologically equivalent doses for LDR-HDR

    International Nuclear Information System (INIS)

    Slosarek, K.; Zajusz, A.

    1996-01-01

    Based on the LQ model examples of biologically equivalent doses LDR, HDR and external beams were calculated. The biologically equivalent doses for LDR were calculated by appending to the LQ model the corrector for the time of repair of radiation sublethal damages. For radiation continuously delivered at a low dose rate the influence of sublethal damage repair time changes on biologically equivalent doses were analysed. For fractionated treatment with high dose rate the biologically equivalent doses were calculated by adding to the LQ model the formula of accelerated repopulation. For total biologically equivalent dose calculation for combine LDR-HDR-Tele irradiation examples are presented with the use of different parameters of the time of repair of sublethal damages and accelerated repopulation. The calculations performed show, that the same biologically equivalent doses can be obtained for different parameters of cell kinetics changes during radiation treatment. It also shows, that during biologically equivalent dose calculations for different radiotherapy schedules, ignorance of cell kinetics parameters can lead to relevant errors

  15. Genitourinary Toxicity After High-Dose-Rate (HDR) Brachytherapy Combined With Hypofractionated External Beam Radiotherapy for Localized Prostate Cancer: An Analysis to Determine the Correlation Between Dose-Volume Histogram Parameters in HDR Brachytherapy and Severity of Toxicity

    International Nuclear Information System (INIS)

    Ishiyama, Hiromichi; Kitano, Masashi; Satoh, Takefumi; Kotani, Shouko; Uemae, Mineko; Matsumoto, Kazumasa; Okusa, Hiroshi; Tabata, Ken-ichi; Baba, Shiro; Hayakawa, Kazushige

    2009-01-01

    Purpose: To evaluate the severity of genitourinary (GU) toxicity in high-dose-rate (HDR) brachytherapy combined with hypofractionated external beam radiotherapy (EBRT) for prostate cancer and to explore factors that might affect the severity of GU toxicity. Methods and Materials: A total of 100 Japanese men with prostate cancer underwent 192 Ir HDR brachytherapy combined with hypofractionated EBRT. Mean (SD) dose to 90% of the planning target volume was 6.3 (0.7) Gy per fraction of HDR. After 5 fractions of HDR treatment, EBRT with 10 fractions of 3 Gy was administrated. The urethral volume receiving 1-15 Gy per fraction in HDR brachytherapy (V1-V15) and the dose to at least 5-100% of urethral volume in HDR brachytherapy (D5-D100) were compared between patients with Grade 3 toxicity and those with Grade 0-2 toxicity. Prostate volume, patient age, and International Prostate Symptom Score were also compared between the two groups. Results: Of the 100 patients, 6 displayed Grade 3 acute GU toxicity, and 12 displayed Grade 3 late GU toxicity. Regarding acute GU toxicity, values of V1, V2, V3, and V4 were significantly higher in patients with Grade 3 toxicity than in those with Grade 0-2 toxicity. Regarding late GU toxicity, values of D70, D80, V12, and V13 were significantly higher in patients with Grade 3 toxicity than in those with Grade 0-2 toxicity. Conclusions: The severity of GU toxicity in HDR brachytherapy combined with hypofractionated EBRT for prostate cancer was relatively high. The volume of prostatic urethra was associated with grade of acute GU toxicity, and urethral dose was associated with grade of late GU toxicity.

  16. Acute genitourinary toxicity after high-dose-rate (HDR) brachytherapy combined with hypofractionated external-beam radiation therapy for localized prostate cancer: Correlation between the urethral dose in HDR brachytherapy and the severity of acute genitourinary toxicity

    International Nuclear Information System (INIS)

    Akimoto, Tetsuo; Ito, Kazuto; Saitoh, Jun-ichi; Noda, Shin-ei; Harashima, Koichi; Sakurai, Hideyuki; Nakayama, Yuko; Yamamoto, Takumi; Suzuki, Kazuhiro; Nakano, Takashi; Niibe, Hideo

    2005-01-01

    Purpose: Several investigations have revealed that the α/β ratio for prostate cancer is atypically low, and that hypofractionation or high-dose-rate (HDR) brachytherapy regimens using appropriate radiation doses may be expected to yield tumor control and late sequelae rates that are better or at least as favorable as those achieved with conventional radiation therapy. In this setting, we attempted treating localized prostate cancer patients with HDR brachytherapy combined with hypofractionated external beam radiation therapy (EBRT). The purpose of this study was to evaluate the feasibility of using this approach, with special emphasis on the relationship between the severity of acute genitourinary (GU) toxicity and the urethral dose calculated from the dose-volume histogram (DVH) of HDR brachytherapy. Methods and Materials: Between September 2000 and December 2003, 70 patients with localized prostate cancer were treated by iridium-192 HDR brachytherapy combined with hypofractionated EBRT at the Gunma University Hospital. Hypofractionated EBRT was administered in fraction doses of 3 Gy, three times per week; a total dose of 51 Gy was delivered to the prostate gland and the seminal vesicles using the four-field technique. No elective pelvic irradiation was performed. After the completion of EBRT, all the patients additionally received transrectal ultrasonography (TRUS)-guided HDR brachytherapy. The fraction size and the number of fractions in HDR brachytherapy were prospectively changed, whereas the total radiation dose for EBRT was fixed at 51 Gy. The fractionation in HDR brachytherapy was as follows: 5 Gy x 5, 7 Gy x 3, 9 Gy x 2, administered twice per day, although the biologic effective dose (BED) for HDR brachytherapy combined with EBRT, assuming that the α/β ratio is 3, was almost equal to 138 in each fractionation group. The planning target volume was defined as the prostate gland with 5-mm margin all around, and the planning was conducted based on

  17. A comparison of anti-tumor effects of high dose rate fractionated and low dose rate continuous irradiation in multicellular spheroids

    International Nuclear Information System (INIS)

    Kubota, Nobuo; Omura, Motoko; Matsubara, Sho.

    1997-01-01

    In a clinical experience, high dose rate (HDR) fractionated interstitial radiotherapy can be an alternative to traditional low dose rate (LDR) continuous interstitial radiotherapy for head and neck cancers. To investigate biological effect of HDR, compared to LDR, comparisons have been made using spheroids of human squamous carcinoma cells. Both LDR and HDR were delivered by 137 Cs at 37degC. Dose rate of LDR was 8 Gy/day and HDR irradiations of fraction size of 4, 5 or 6 Gy were applied twice a day with an interval time of more than 6 hr. We estimated HDR fractionated dose of 31 Gy with 4 Gy/fr to give the same biological effects of 38 Gy by continuous LDR for spheroids. The ratio of HDR/LDR doses to control 50% spheroids was 0.82. (author)

  18. The optimal fraction size in high-dose-rate brachytherapy: dependency on tissue repair kinetics and low-dose rate

    International Nuclear Information System (INIS)

    Sminia, Peter; Schneider, Christoph J.; Fowler, Jack F.

    2002-01-01

    Background and Purpose: Indications of the existence of long repair half-times on the order of 2-4 h for late-responding human normal tissues have been obtained from continuous hyperfractionated accelerated radiotherapy (CHART). Recently, these data were used to explain, on the basis of the biologically effective dose (BED), the potential superiority of fractionated high-dose rate (HDR) with large fraction sizes of 5-7 Gy over continuous low-dose rate (LDR) irradiation at 0.5 Gy/h in cervical carcinoma. We investigated the optimal fraction size in HDR brachytherapy and its dependency on treatment choices (overall treatment time, number of HDR fractions, and time interval between fractions) and treatment conditions (reference low-dose rate, tissue repair characteristics). Methods and Materials: Radiobiologic model calculations were performed using the linear-quadratic model for incomplete mono-exponential repair. An irradiation dose of 20 Gy was assumed to be applied either with HDR in 2-12 fractions or continuously with LDR for a range of dose rates. HDR and LDR treatment regimens were compared on the basis of the BED and BED ratio of normal tissue and tumor, assuming repair half-times between 1 h and 4 h. Results: With the assumption that the repair half-time of normal tissue was three times longer than that of the tumor, hypofractionation in HDR relative to LDR could result in relative normal tissue sparing if the optimum fraction size is selected. By dose reduction while keeping the tumor BED constant, absolute normal tissue sparing might therefore be achieved. This optimum HDR fraction size was found to be largely dependent on the LDR dose rate. On the basis of the BED NT/TUM ratio of HDR over LDR, 3 x 6.7 Gy would be the optimal HDR fractionation scheme for replacement of an LDR scheme of 20 Gy in 10-30 h (dose rate 2-0.67 Gy/h), while at a lower dose rate of 0.5 Gy/h, four fractions of 5 Gy would be preferential, still assuming large differences between tumor

  19. Dose and dose rate effects of whole-body gamma-irradiation: I. Lymphocytes and lymphoid organs

    Science.gov (United States)

    Pecaut, M. J.; Nelson, G. A.; Gridley, D. S.

    2001-01-01

    The major goal of part I of this study was to compare varying doses and dose rates of whole-body gamma-radiation on lymphoid cells and organs. C57BL/6 mice (n = 75) were exposed to 0, 0.5, 1.5, and 3.0 Gy gamma-rays (60Co) at 1 cGy/min (low-dose rate, LDR) and 80 cGy/min (high-dose rate, HDR) and euthanized 4 days later. A significant dose-dependent loss of spleen mass was observed with both LDR and HDR irradiation; for the thymus this was true only with HDR. Decreasing leukocyte and lymphocyte numbers occurred with increasing dose in blood and spleen at both dose rates. The numbers (not percentages) of CD3+ T lymphocytes decreased in the blood in a dose-dependent manner at both HDR and LDR. Splenic T cell counts decreased with dose only in HDR groups; percentages increased with dose at both dose rates. Dose-dependent decreases occurred in CD4+ T helper and CD8+ T cytotoxic cell counts at HDR and LDR. In the blood the percentages of CD4+ cells increased with increasing dose at both dose rates, whereas in the spleen the counts decreased only in the HDR groups. The percentages of the CD8+ population remained stable in both blood and spleen. CD19+ B cell counts and percentages in both compartments declined markedly with increasing HDR and LDR radiation. NK1.1+ natural killer cell numbers and proportions remained relatively stable. Overall, these data indicate that the observed changes were highly dependent on the dose, but not dose rate, and that cells in the spleen are more affected by dose rate than those in blood. The results also suggest that the response of lymphocytes in different body compartments may be variable.

  20. High dose rate versus low dose rate interstitial radiotherapy for carcinoma of the floor of mouth

    International Nuclear Information System (INIS)

    Inoue, Takehiro; Inoue, Toshihiko; Yamazaki, Hideya; Koizumi, Masahiko; Kagawa, Kazufumi; Yoshida, Ken; Shiomi, Hiroya; Imai, Atsushi; Shimizutani, Kimishige; Tanaka, Eichii; Nose, Takayuki; Teshima, Teruki; Furukawa, Souhei; Fuchihata, Hajime

    1998-01-01

    Purpose: Patients with cancer of the floor of mouth are treated with radiation because of functional and cosmetic reasons. We evaluate the treatment results of high dose rate (HDR) and low dose rate (LDR) interstitial radiation for cancer of the floor of mouth. Methods and Materials: From January 1980 through March 1996, 41 patients with cancer of the floor of mouth were treated with LDR interstitial radiation using 198 Au grains, and from April 1992 through March 1996 16 patients with HDR interstitial radiation. There were 26 T1 tumors, 30 T2 tumors, and 1 T3 tumor. For 21 patients treated with interstitial radiation alone, a total radiation dose of interstitial therapy was 60 Gy/10 fractions/6-7 days in HDR and 85 Gy within 1 week in LDR. For 36 patients treated with a combination therapy, a total dose of 30 to 40 Gy of external radiation and a total dose of 48 Gy/8 fractions/5-6 days in HDR or 65 Gy within 1 week in LDR were delivered. Results: Two- and 5-year local control rates of patients treated with HDR interstitial radiation were 94% and 94%, and those with LDR were 75% and 69%, respectively. Local control rate of patients treated with HDR brachytherapy was slightly higher than that with 198 Au grains (p = 0.113). For late complication, bone exposure or ulcer occurred in 6 of 16 (38%) patients treated with HDR and 13 of 41 (32%) patients treated with LDR. Conclusion: HDR fractionated interstitial brachytherapy can be an alternative to LDR brachytherapy for cancer of the floor of mouth and eliminate radiation exposure for the medical staff

  1. High Dose-Rate Versus Low Dose-Rate Brachytherapy for Lip Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Ghadjar, Pirus, E-mail: pirus.ghadjar@insel.ch [Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern (Switzerland); Bojaxhiu, Beat [Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern (Switzerland); Simcock, Mathew [Swiss Group for Clinical Cancer Research Coordinating Center, Bern (Switzerland); Terribilini, Dario; Isaak, Bernhard [Division of Medical Radiation Physics, Inselspital, Bern University Hospital, and University of Bern, Bern (Switzerland); Gut, Philipp; Wolfensberger, Patrick; Broemme, Jens O.; Geretschlaeger, Andreas; Behrensmeier, Frank; Pica, Alessia; Aebersold, Daniel M. [Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern (Switzerland)

    2012-07-15

    Purpose: To analyze the outcome after low-dose-rate (LDR) or high-dose-rate (HDR) brachytherapy for lip cancer. Methods and Materials: One hundred and three patients with newly diagnosed squamous cell carcinoma of the lip were treated between March 1985 and June 2009 either by HDR (n = 33) or LDR brachytherapy (n = 70). Sixty-eight patients received brachytherapy alone, and 35 received tumor excision followed by brachytherapy because of positive resection margins. Acute and late toxicity was assessed according to the Common Terminology Criteria for Adverse Events 3.0. Results: Median follow-up was 3.1 years (range, 0.3-23 years). Clinical and pathological variables did not differ significantly between groups. At 5 years, local recurrence-free survival, regional recurrence-free survival, and overall survival rates were 93%, 90%, and 77%. There was no significant difference for these endpoints when HDR was compared with LDR brachytherapy. Forty-two of 103 patients (41%) experienced acute Grade 2 and 57 of 103 patients (55%) experienced acute Grade 3 toxicity. Late Grade 1 toxicity was experienced by 34 of 103 patients (33%), and 5 of 103 patients (5%) experienced late Grade 2 toxicity; no Grade 3 late toxicity was observed. Acute and late toxicity rates were not significantly different between HDR and LDR brachytherapy. Conclusions: As treatment for lip cancer, HDR and LDR brachytherapy have comparable locoregional control and acute and late toxicity rates. HDR brachytherapy for lip cancer seems to be an effective treatment with acceptable toxicity.

  2. High Dose-Rate Versus Low Dose-Rate Brachytherapy for Lip Cancer

    International Nuclear Information System (INIS)

    Ghadjar, Pirus; Bojaxhiu, Beat; Simcock, Mathew; Terribilini, Dario; Isaak, Bernhard; Gut, Philipp; Wolfensberger, Patrick; Brömme, Jens O.; Geretschläger, Andreas; Behrensmeier, Frank; Pica, Alessia; Aebersold, Daniel M.

    2012-01-01

    Purpose: To analyze the outcome after low-dose-rate (LDR) or high-dose-rate (HDR) brachytherapy for lip cancer. Methods and Materials: One hundred and three patients with newly diagnosed squamous cell carcinoma of the lip were treated between March 1985 and June 2009 either by HDR (n = 33) or LDR brachytherapy (n = 70). Sixty-eight patients received brachytherapy alone, and 35 received tumor excision followed by brachytherapy because of positive resection margins. Acute and late toxicity was assessed according to the Common Terminology Criteria for Adverse Events 3.0. Results: Median follow-up was 3.1 years (range, 0.3–23 years). Clinical and pathological variables did not differ significantly between groups. At 5 years, local recurrence-free survival, regional recurrence-free survival, and overall survival rates were 93%, 90%, and 77%. There was no significant difference for these endpoints when HDR was compared with LDR brachytherapy. Forty-two of 103 patients (41%) experienced acute Grade 2 and 57 of 103 patients (55%) experienced acute Grade 3 toxicity. Late Grade 1 toxicity was experienced by 34 of 103 patients (33%), and 5 of 103 patients (5%) experienced late Grade 2 toxicity; no Grade 3 late toxicity was observed. Acute and late toxicity rates were not significantly different between HDR and LDR brachytherapy. Conclusions: As treatment for lip cancer, HDR and LDR brachytherapy have comparable locoregional control and acute and late toxicity rates. HDR brachytherapy for lip cancer seems to be an effective treatment with acceptable toxicity.

  3. Effect and toxicity of endoluminal high-dose-rate (HDR) brachytherapy in centrally located tumors of the upper respiratory tract

    International Nuclear Information System (INIS)

    Harms, W.; Wannenmacher, M.; Becker, H.; Herth, F.; Fritz, P.

    2000-01-01

    Aim: To assess effect an toxicity of high-dose-rate afterloading (HDR) alone or in combination with external beam radiotherapy (EBRT) in centrally located tumors of the upper respiratory tract. Patients and Methods: From 1987 to 1996, 55 patients were treated. Twenty-one patients (group A1: 17 non-small-cell lung cancer [NSCLC], A2: 4 metastases from other malignancies) were treated using HDR alone due to a relapse after external beam irradiation. In 34 previously untreated and inoperable patients (group B1: 27 NSCLC, B2: 7 metastases from other malignancies) HDR was given as a boost after EBRT (30 to 60 Gy, median 50). HDR was carried out with a 192 Ir source (370 GBq). The brachytherapy dose (group A: 5 to 27 Gy, median 20; B: 10 to 20 Gy, median 15) was prescribed to 1 cm distance from the source axis. A distanciable applicator was used in 39/55 patients. Results: In group A1, a response rate (CR, PR) of 53% (group B1: 77%) was reached. The median survival (Kaplan-Meier) was 5 months in group A1 (B1: 20 months). The 1-, 3- and 5-year local progression free survival rates (Kaplan-Meier) were 66% (15%), 52% (0%), and 37% (0%) in group B1 (group A1). Prognostic favorable factors in group B1 were a tumor diameter 70. Grade-1 or 2 toxicity (RTOG/EORTC) occurred in 0% in group A and in 6% in group B. We observed no Grad-3 or 4 toxicity. Complications caused by persistent or progressive local disease occurred in 3 patients in goup A (fatal hemorrhage, tracheomediastinal fistula, hemoptysis) and in 2 patients in group B (fatal hemorrhage, hemoptysis). Conclusions: HDR brachytherapy is an effective treatment with moderate side effects. In combination with external beam irradiation long-term remissions can be reached in one third of the patients. (orig.) [de

  4. Comparison of traditional low-dose-rate to optimized and nonoptimized high-dose-rate tandem and ovoid dosimetry

    International Nuclear Information System (INIS)

    Decker, William E.; Erickson, Beth; Albano, Katherine; Gillin, Michael

    2001-01-01

    Purpose: Few dose specification guidelines exist when attempting to perform high-dose-rate (HDR) dosimetry. The purpose of this study was to model low-dose-rate (LDR) dosimetry, using parameters common in HDR dosimetry, to achieve the 'pear-shape' dose distribution achieved with LDR tandem and ovoid applications. Methods and Materials: Radiographs of Fletcher-Suit LDR applicators and Nucletron 'Fletcher-like' HDR applicators were taken with the applicators in an idealized geometry. Traditional Fletcher loadings of 3M Cs-137 sources and the Theratronics Planning System were used for LDR dosimetry. HDR dosimetry was performed using the Nucletron Microselectron HDR UPS V11.22 with an Ir-192 source. Dose optimization points were initially located along a line 2 cm lateral to the tandem, beginning at the tandem tip at 0.5-cm intervals, ending at the sail, and optimized to 100% of the point A dose. A single dose optimization point was also placed laterally from the center of each ovoid equal to the radius of the ovoid (ovoid surface dose). For purposes of comparison, dose was also calculated for points A and B, and a point located 1 cm superior to the tandem tip in the plane of the tandem, (point F). Four- and 6-cm tandem lengths and 2.0-, 2.5-, and 3.0-cm ovoid diameters were used for this study. Based on initial findings, dose optimization schemes were developed to best approximate LDR dosimetry. Finally, radiographs were obtained of HDR applications in two patients. These radiographs were used to compare the optimization schemes with 'nonoptimized' treatment plans. Results: Calculated doses for points A and B were similar for LDR, optimized HDR, and nonoptimized HDR. The optimization scheme that used tapered dose points at the tandem tip and optimized a single ovoid surface point on each ovoid to 170% of point A resulted in a good approximation of LDR dosimetry. Nonoptimized HDR resulted in higher doses at point F, the bladder, and at points lateral to the tandem tip

  5. Inverse dose-rate-effects on the expressions of extra-cellular matrix-related genes in low-dose-rate γ-ray irradiated murine cells

    International Nuclear Information System (INIS)

    Sugihara, Takashi; Tanaka, Kimio; Oghiso, Yoichi; Murano, Hayato

    2008-01-01

    Based on the results of previous microarray analyses of murine NIH3T3/PG13Luc cells irradiated with continuous low-dose-rate (LDR) γ-ray or end-high-dose-rate-irradiations (end-HDR) at the end of the LDR-irradiation period, the inverse dose-rate-effects on gene expression levels were observed. To compare differences of the effects between LDR-irradiation and HDR-irradiation, HDR-irradiations at 2 different times, one (ini-HDR) at the same time at the start of LDR-irradiation and the other (end-HDR), were performed. The up-regulated genes were classified into two types, in which one was up-regulated in LDR-, ini-HDR-, and end-HDR irradiation such as Cdkn1a and Ccng1, which were reported as p53-dependent genes, and the other was up-regulated in LDR- and ini-HDR irradiations such as pro-collagen TypeIa2/Colla2, TenascinC/Tnc, and Fibulin5/Fbln5, which were reported as extra-cellular matrix-related (ECM) genes. The time dependent gene expression patterns in LDR-irradiation were also classified into two types, in which one was an early response such as in Cdkn1a and Ccng1 and the other was a delayed response such as the ECM genes which have no linearity to total dose. The protein expression pattern of Cdkn1a increased dose dependently in LDR- and end-HDR-irradiations, but those of p53Ser15/18 and MDM2 in LDR-irradiations were different from end-HDR-irradiations. Furthermore, the gene expression levels of the ECM genes in embryonic fibroblasts from p53-deficient mice were not increased by LDR- and end-HDR-irradiation, so the delayed expressions of the ECM genes seem to be regulated by p53. Consequently, the inverse dose-rate-effects on the expression levels of the ECM genes in LDR- and end-HDR-irradiations may be explained from different time responses by p53 status. (author)

  6. Quality control of 192Ir high dose rate after loading brachytherapy dose veracity

    International Nuclear Information System (INIS)

    Feng Zhongsu; Xu Xiao; Liu Fen

    2008-01-01

    Recently, 192 Ir high dose rate (HDR) afterloading are widely used in brachytherapy. The advantage of using HDR systems over low dose rate systems are shorter treatment time and higher fraction dose. To guarantee the veracity of the delivery dose, several quality control methods are deseribed in this work. With these we can improve the position precision, time precision and dose precision of the brachytherapy. (authors)

  7. Radiation Parameters of High Dose Rate Iridium -192 Sources

    Science.gov (United States)

    Podgorsak, Matthew B.

    A lack of physical data for high dose rate (HDR) Ir-192 sources has necessitated the use of basic radiation parameters measured with low dose rate (LDR) Ir-192 seeds and ribbons in HDR dosimetry calculations. A rigorous examination of the radiation parameters of several HDR Ir-192 sources has shown that this extension of physical data from LDR to HDR Ir-192 may be inaccurate. Uncertainty in any of the basic radiation parameters used in dosimetry calculations compromises the accuracy of the calculated dose distribution and the subsequent dose delivery. Dose errors of up to 0.3%, 6%, and 2% can result from the use of currently accepted values for the half-life, exposure rate constant, and dose buildup effect, respectively. Since an accuracy of 5% in the delivered dose is essential to prevent severe complications or tumor regrowth, the use of basic physical constants with uncertainties approaching 6% is unacceptable. A systematic evaluation of the pertinent radiation parameters contributes to a reduction in the overall uncertainty in HDR Ir-192 dose delivery. Moreover, the results of the studies described in this thesis contribute significantly to the establishment of standardized numerical values to be used in HDR Ir-192 dosimetry calculations.

  8. Comparison of high-dose-rate and low-dose-rate brachytherapy in the treatment of endometrial carcinoma

    International Nuclear Information System (INIS)

    Fayed, Alaa; Mutch, David G.; Rader, Janet S.; Gibb, Randall K.; Powell, Matthew A.; Wright, Jason D.; El Naqa, Issam; Zoberi, Imran; Grigsby, Perry W.

    2007-01-01

    Purpose: To compare the outcomes for endometrial carcinoma patients treated with either high-dose-rate (HDR) or low-dose-rate (LDR) brachytherapy. Methods and Materials: This study included 1,179 patients divided into LDR (1,004) and HDR groups (175). Patients with International Federation of Gynecology and Obstetrics (FIGO) surgical Stages I-III were included. All patients were treated with postoperative irradiation. In the LDR group, the postoperative dose applied to the vaginal cuff was 60-70 Gy surface doses to the vaginal mucosa. The HDR brachytherapy prescription was 6 fractions of 2 Gy each to a depth of 0.5 cm from the surface of the vaginal mucosa. Overall survival, disease-free survival, local control, and complications were endpoints. Results: For all stages combined, the overall survival, disease-free survival, and local control at 5 years in the LDR group were 70%, 69%, and 81%, respectively. For all stages combined, the overall survival, disease-free survival, and local control at 5 years in the HDR group were 68%, 62%, and 78%, respectively. There were no significant differences in early or late Grade III and IV complications in the HDR or LDR groups. Conclusion: Survival outcomes, pelvic tumor control, and Grade III and IV complications were not significantly different in the LDR brachytherapy group compared with the HDR group

  9. High dose rate versus medium dose rate intraluminal brachytherapy in inoperable esophageal carcinoma

    International Nuclear Information System (INIS)

    Langendijk, J.; Jager, J.; Jong, J. de; Rijken, J.; Pannebakker, M.

    1996-01-01

    Introduction: The purpose of this study was to compare the results of medium dose rate (MDR) intraluminal brachytherapy (ILBT) and high dose rate (HDR) ILBT in patients with inoperable esophageal carcinoma, with regard to dysphagia, complication rate and survival. Material and methods: Included were 114 patients with inoperable esophageal cancer who were treated with a single session of ILBT. In all cases a single dose of 15 Gy was administered, calculated at a 1 cm radius. Forty-eight patients were treated with MDR ( 137 Cs)ILBT. In June 1990 MDR was replaced by HDR and from then 66 patients were treated with HDR ( 192 Ir). Dysphagia was prospectively scored using a 5-point scale at 6 weeks, 3, 6, 9 and 12 months. Results: No significant differences were noted between the two groups with regard to pretreatment variables. In patients treated with MDR-ILBT improvement of swallowing ability was noted in 30 out of 42 evaluable patients (71%), no change in 9 (21%) and progression of dysphagia in 3 patients (8%), as compared to 34 out of 59 evaluable patients (58%), 16 (27%) and 6 (15%) resp. in de HDR-ILBT group. In the latter category, progression of dysphagia was caused by fistulae in 2 patients. The differences were not significant (ns). Additional treatment in case of recurrent or persistent dysphagia was needed in 50% of the cases in the MDR-ILBT group as compared to 41% in the HDR-ILBT group (ns). The median survival of the MDR-ILBT group was 3.9 months as compared to 4.3 months in the HDR-ILBT group (ns). In 2 patients (4%) treated with MDR-ILBT bronchio-oesphageal fistulae developed at 6 weeks and 2 months. In the HDR-ILBT group fistulae were noted in 7 cases (11%) at 2 weeks, 4 weeks, 2, 3, 3, 4 and 9 months (ns). In all of these cases persistent of recurrent tumour was present. Conclusions: No significant differences were noted with regard to palliation of dysphagia, survival and complication rate between MDR-ILBT and HDR-ILBT in the management of esophageal

  10. Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (HDR) brachytherapy

    Science.gov (United States)

    Palmer, Antony L.; Lee, Chris; Ratcliffe, Ailsa J.; Bradley, David; Nisbet, Andrew

    2013-10-01

    A novel phantom is presented for ‘full system’ dosimetric audit comparing planned and delivered dose distributions in HDR gynaecological brachytherapy, using clinical treatment applicators. The brachytherapy applicator dosimetry test object consists of a near full-scatter water tank with applicator and film supports constructed of Solid Water, accommodating any typical cervix applicator. Film dosimeters are precisely held in four orthogonal planes bisecting the intrauterine tube, sampling dose distributions in the high risk clinical target volume, points A and B, bladder, rectum and sigmoid. The applicator position is fixed prior to CT scanning and through treatment planning and irradiation. The CT data is acquired with the applicator in a near clinical orientation to include applicator reconstruction in the system test. Gamma analysis is used to compare treatment planning system exported RTDose grid with measured multi-channel film dose maps. Results from two pilot audits are presented, using Ir-192 and Co-60 HDR sources, with a mean gamma passing rate of 98.6% using criteria of 3% local normalization and 3 mm distance to agreement (DTA). The mean DTA between prescribed dose and measured film dose at point A was 1.2 mm. The phantom was funded by IPEM and will be used for a UK national brachytherapy dosimetry audit.

  11. Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (HDR) brachytherapy

    International Nuclear Information System (INIS)

    Palmer, Antony L; Bradley, David; Nisbet, Andrew; Lee, Chris; Ratcliffe, Ailsa J

    2013-01-01

    A novel phantom is presented for ‘full system’ dosimetric audit comparing planned and delivered dose distributions in HDR gynaecological brachytherapy, using clinical treatment applicators. The brachytherapy applicator dosimetry test object consists of a near full-scatter water tank with applicator and film supports constructed of Solid Water, accommodating any typical cervix applicator. Film dosimeters are precisely held in four orthogonal planes bisecting the intrauterine tube, sampling dose distributions in the high risk clinical target volume, points A and B, bladder, rectum and sigmoid. The applicator position is fixed prior to CT scanning and through treatment planning and irradiation. The CT data is acquired with the applicator in a near clinical orientation to include applicator reconstruction in the system test. Gamma analysis is used to compare treatment planning system exported RTDose grid with measured multi-channel film dose maps. Results from two pilot audits are presented, using Ir-192 and Co-60 HDR sources, with a mean gamma passing rate of 98.6% using criteria of 3% local normalization and 3 mm distance to agreement (DTA). The mean DTA between prescribed dose and measured film dose at point A was 1.2 mm. The phantom was funded by IPEM and will be used for a UK national brachytherapy dosimetry audit. (paper)

  12. Acute genitourinary toxicity after high dose rate (HDR) brachytherapy combined with hypofractionated external-beam radiation therapy for localized prostate cancer: Second analysis to determine the correlation between the urethral dose in HDR brachytherapy and the severity of acute genitourinary toxicity

    International Nuclear Information System (INIS)

    Akimoto, Tetsuo; Katoh, Hiroyuki; Noda, Shin-ei; Ito, Kazuto; Yamamoto, Takumi; Kashiwagi, Bunzo; Nakano, Takashi

    2005-01-01

    Purpose: We have been treating localized prostate cancer with high-dose-rate (HDR) brachytherapy combined with hypofractionated external beam radiation therapy (EBRT) at our institution. We recently reported the existence of a correlation between the severity of acute genitourinary (GU) toxicity and the urethral radiation dose in HDR brachytherapy by using different fractionation schema. The purpose of this study was to evaluate the role of the urethral dose in the development of acute GU toxicity more closely than in previous studies. For this purpose, we conducted an analysis of patients who had undergone HDR brachytherapy with a fixed fractionation schema combined with hypofractionated EBRT. Methods and Materials: Among the patients with localized prostate cancer who were treated by 192-iridium HDR brachytherapy combined with hypofractionated EBRT at Gunma University Hospital between August 2000 and November 2004, we analyzed 67 patients who were treated by HDR brachytherapy with the fractionation schema of 9 Gy x two times combined with hypofractionated EBRT. Hypofractionated EBRT was administered at a fraction dose of 3 Gy three times weekly, and a total dose of 51 Gy was delivered to the prostate gland and seminal vesicles using the four-field technique. No elective pelvic irradiation was performed. After the completion of EBRT, all the patients additionally received transrectal ultrasonography-guided HDR brachytherapy. The planning target volume was defined as the prostate gland with a 5-mm margin all around, and the planning was conducted based on computed tomography images. The tumor stage was T1c in 13 patients, T2 in 31 patients, and T3 in 23 patients. The Gleason score was 2-6 in 12 patients, 7 in 34 patients, and 8-10 in 21 patients. Androgen ablation was performed in all the patients. The median follow-up duration was 11 months (range 3-24 months). The toxicities were graded based on the Radiation Therapy Oncology Group and the European Organization

  13. High dose rate brachytherapy for oral cancer.

    Science.gov (United States)

    Yamazaki, Hideya; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Furukawa, Souhei; Koizumi, Masahiko; Ogawa, Kazuhiko

    2013-01-01

    Brachytherapy results in better dose distribution compared with other treatments because of steep dose reduction in the surrounding normal tissues. Excellent local control rates and acceptable side effects have been demonstrated with brachytherapy as a sole treatment modality, a postoperative method, and a method of reirradiation. Low-dose-rate (LDR) brachytherapy has been employed worldwide for its superior outcome. With the advent of technology, high-dose-rate (HDR) brachytherapy has enabled health care providers to avoid radiation exposure. This therapy has been used for treating many types of cancer such as gynecological cancer, breast cancer, and prostate cancer. However, LDR and pulsed-dose-rate interstitial brachytherapies have been mainstays for head and neck cancer. HDR brachytherapy has not become widely used in the radiotherapy community for treating head and neck cancer because of lack of experience and biological concerns. On the other hand, because HDR brachytherapy is less time-consuming, treatment can occasionally be administered on an outpatient basis. For the convenience and safety of patients and medical staff, HDR brachytherapy should be explored. To enhance the role of this therapy in treatment of head and neck lesions, we have reviewed its outcomes with oral cancer, including Phase I/II to Phase III studies, evaluating this technique in terms of safety and efficacy. In particular, our studies have shown that superficial tumors can be treated using a non-invasive mold technique on an outpatient basis without adverse reactions. The next generation of image-guided brachytherapy using HDR has been discussed. In conclusion, although concrete evidence is yet to be produced with a sophisticated study in a reproducible manner, HDR brachytherapy remains an important option for treatment of oral cancer.

  14. Pulsed dose rate and fractionated high dose rate brachytherapy: choice of brachytherapy schedules to replace low dose rate treatments

    International Nuclear Information System (INIS)

    Visser, Andries G.; Aardweg, Gerard J.M.J. van den; Levendag, Peter C.

    1996-01-01

    Purpose: Pulsed dose rate (PDR) brachytherapy is a new type of afterloading brachytherapy (BT) in which a continuous low dose rate (LDR) treatment is simulated by a series of 'pulses,' i.e., fractions of short duration (less than 0.5 h) with intervals between fractions of 1 to a few hours. At the Dr. Daniel den Hoed Cancer Center, the term 'PDR brachytherapy' is used for treatment schedules with a large number of fractions (at least four per day), while the term 'fractionated high dose rate (HDR) brachytherapy' is used for treatment schedules with just one or two brachytherapy fractions per day. Both treatments can be applied as alternatives for LDR BT. This article deals with the choice between PDR and fractionated HDR schedules and proposes possible fractionation schedules. Methods and Materials: To calculate HDR and PDR fractionation schedules with the intention of being equivalent to LDR BT, the linear-quadratic (LQ) model has been used in an incomplete repair formulation as given by Brenner and Hall, and by Thames. In contrast to earlier applications of this model, both the total physical dose and the overall time were not kept identical for LDR and HDR/PDR schedules. A range of possible PDR treatment schedules is presented, both for booster applications (in combination with external radiotherapy (ERT) and for BT applications as a single treatment. Because the knowledge of both α/β values and the half time for repair of sublethal damage (T (1(2)) ), which are required for these calculations, is quite limited, calculations regarding the equivalence of LDR and PDR treatments have been performed for a wide range of values of α/β and T (1(2)) . The results are presented graphically as PDR/LDR dose ratios and as ratios of the PDR/LDR tumor control probabilities. Results: If the condition that total physical dose and overall time of a PDR treatment must be exactly identical to the values for the corresponding LDR treatment regimen is not applied, there appears

  15. Cost minimization analysis of high-dose-rate versus low-dose-rate brachytherapy in endometrial cancer

    International Nuclear Information System (INIS)

    Pinilla, James

    1998-01-01

    Purpose: Endometrial cancer is a common, usually curable malignancy whose treatment frequently involves low-dose-rate (LDR) or high-dose-rate (HDR) brachytherapy. These treatments involve substantial resource commitments and this is increasingly important. This paper presents a cost minimization analysis of HDR versus LDR brachytherapy in the treatment of endometrial cancer. Methods and Materials: The perspective of the analysis is that of the payor, in this case the Ministry of Health. One course of LDR treatment is compared to two courses of HDR treatment. The two alternatives are considered to be comparable with respect to local control, survival, and toxicities. Labor, overhead, and capital costs are accounted for and carefully measured. A 5% inflation rate is used where applicable. A univariate sensitivity analysis is performed. Results: The HDR regime is 22% less expensive compared to the LDR regime. This is $991.66 per patient or, based on the current workload of this department (30 patients per year) over the useful lifetime of the after loader, $297,498 over 10 years in 1997 dollars. Conclusion: HDR brachytherapy minimizes costs in the treatment of endometrial cancer relative to LDR brachytherapy. These results may be used by other centers to make rational decisions regarding brachytherapy equipment replacement or acquisition

  16. High and low dose-rate brachytherapy for cervical carcinoma

    International Nuclear Information System (INIS)

    Orton, C.G.

    1998-01-01

    For the brachytherapy component of the r[iation treatment of cervical carcinoma, high dose rate (HDR) is slowly replacing conventional low dose rate (LDR) due primarily to r[iation safety and other physical benefits attributed to the HDR modality. Many r[iation oncologists are reluctant to make this change because of perceived r[iobiological dis[vantages of HDR. However, in clinical practice HDR appears to be as effective as LDR but with a lower risk of late complications, as demonstrated by one randomized clinical trial and two comprehensive literature and practice surveys. The reason for this appears to be that the r[iobiological dis[vantages of HDR are outweighed by the physical [vantages. (orig.)

  17. Determination of the dose of traffic in HDR brachytherapy with ALANINE/R PE technique

    International Nuclear Information System (INIS)

    Guzman Calcina, C. S.; Chen, F.; Almeida, A. de; Baffa, O.

    2001-01-01

    It determines, experimentally, the dose of traffic in brachytherapy for High Dose Rate (HDR), using for the first-time the Electronic Paramagnetic Resonance (EPR) technique with alanine detectors. The value obtained is the published next to obtained using lithium fluoride thermoluminescent dosimeters [es

  18. A segmentation and point-matching enhanced efficient deformable image registration method for dose accumulation between HDR CT images

    International Nuclear Information System (INIS)

    Zhen, Xin; Chen, Haibin; Zhou, Linghong; Yan, Hao; Jiang, Steve; Jia, Xun; Gu, Xuejun; Mell, Loren K; Yashar, Catheryn M; Cervino, Laura

    2015-01-01

    Deformable image registration (DIR) of fractional high-dose-rate (HDR) CT images is challenging due to the presence of applicators in the brachytherapy image. Point-to-point correspondence fails because of the undesired deformation vector fields (DVF) propagated from the applicator region (AR) to the surrounding tissues, which can potentially introduce significant DIR errors in dose mapping. This paper proposes a novel segmentation and point-matching enhanced efficient DIR (named SPEED) scheme to facilitate dose accumulation among HDR treatment fractions. In SPEED, a semi-automatic seed point generation approach is developed to obtain the incremented fore/background point sets to feed the random walks algorithm, which is used to segment and remove the AR, leaving empty AR cavities in the HDR CT images. A feature-based ‘thin-plate-spline robust point matching’ algorithm is then employed for AR cavity surface points matching. With the resulting mapping, a DVF defining on each voxel is estimated by B-spline approximation, which serves as the initial DVF for the subsequent Demons-based DIR between the AR-free HDR CT images. The calculated DVF via Demons combined with the initial one serve as the final DVF to map doses between HDR fractions. The segmentation and registration accuracy are quantitatively assessed by nine clinical HDR cases from three gynecological cancer patients. The quantitative analysis and visual inspection of the DIR results indicate that SPEED can suppress the impact of applicator on DIR, and accurately register HDR CT images as well as deform and add interfractional HDR doses. (paper)

  19. A segmentation and point-matching enhanced efficient deformable image registration method for dose accumulation between HDR CT images

    Science.gov (United States)

    Zhen, Xin; Chen, Haibin; Yan, Hao; Zhou, Linghong; Mell, Loren K.; Yashar, Catheryn M.; Jiang, Steve; Jia, Xun; Gu, Xuejun; Cervino, Laura

    2015-04-01

    Deformable image registration (DIR) of fractional high-dose-rate (HDR) CT images is challenging due to the presence of applicators in the brachytherapy image. Point-to-point correspondence fails because of the undesired deformation vector fields (DVF) propagated from the applicator region (AR) to the surrounding tissues, which can potentially introduce significant DIR errors in dose mapping. This paper proposes a novel segmentation and point-matching enhanced efficient DIR (named SPEED) scheme to facilitate dose accumulation among HDR treatment fractions. In SPEED, a semi-automatic seed point generation approach is developed to obtain the incremented fore/background point sets to feed the random walks algorithm, which is used to segment and remove the AR, leaving empty AR cavities in the HDR CT images. A feature-based ‘thin-plate-spline robust point matching’ algorithm is then employed for AR cavity surface points matching. With the resulting mapping, a DVF defining on each voxel is estimated by B-spline approximation, which serves as the initial DVF for the subsequent Demons-based DIR between the AR-free HDR CT images. The calculated DVF via Demons combined with the initial one serve as the final DVF to map doses between HDR fractions. The segmentation and registration accuracy are quantitatively assessed by nine clinical HDR cases from three gynecological cancer patients. The quantitative analysis and visual inspection of the DIR results indicate that SPEED can suppress the impact of applicator on DIR, and accurately register HDR CT images as well as deform and add interfractional HDR doses.

  20. Objective method to report planner-independent skin/rib maximal dose in balloon-based high dose rate (HDR) brachytherapy for breast cancer

    International Nuclear Information System (INIS)

    Kim, Yongbok; Trombetta, Mark G.

    2011-01-01

    Purpose: An objective method was proposed and compared with a manual selection method to determine planner-independent skin and rib maximal dose in balloon-based high dose rate (HDR) brachytherapy planning. Methods: The maximal dose to skin and rib was objectively extracted from a dose volume histogram (DVH) of skin and rib volumes. A virtual skin volume was produced by expanding the skin surface in three dimensions (3D) external to the breast with a certain thickness in the planning computed tomography (CT) images. Therefore, the maximal dose to this volume occurs on the skin surface the same with a conventional manual selection method. The rib was also delineated in the planning CT images and its maximal dose was extracted from its DVH. The absolute (Abdiff=|D max Man -D max DVH |) and relative (Rediff[%]=100x(|D max Man -D max DVH |)/D max DVH ) maximal skin and rib dose differences between the manual selection method (D max Man ) and the objective method (D max DVH ) were measured for 50 balloon-based HDR (25 MammoSite and 25 Contura) patients. Results: The average±standard deviation of maximal dose difference was 1.67%±1.69% of the prescribed dose (PD). No statistical difference was observed between MammoSite and Contura patients for both Abdiff and Rediff[%] values. However, a statistically significant difference (p value max >90%) compared with lower dose range (D max <90%): 2.16%±1.93% vs 1.19%±1.25% with p value of 0.0049. However, the Rediff[%] analysis eliminated the inverse square factor and there was no statistically significant difference (p value=0.8931) between high and low dose ranges. Conclusions: The objective method using volumetric information of skin and rib can determine the planner-independent maximal dose compared with the manual selection method. However, the difference was <2% of PD, on average, if appropriate attention is paid to selecting a manual dose point in 3D planning CT images.

  1. High dose rate brachytherapy for oral cancer

    International Nuclear Information System (INIS)

    Yamazaki, Hideya; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Koizumi, Masahiko; Ogawa, Kazuhiko; Furukawa, Souhei

    2013-01-01

    Brachytherapy results in better dose distribution compared with other treatments because of steep dose reduction in the surrounding normal tissues. Excellent local control rates and acceptable side effects have been demonstrated with brachytherapy as a sole treatment modality, a postoperative method, and a method of reirradiation. Low-dose-rate (LDR) brachytherapy has been employed worldwide for its superior outcome. With the advent of technology, high-dose-rate (HDR) brachytherapy has enabled health care providers to avoid radiation exposure. This therapy has been used for treating many types of cancer such as gynecological cancer, breast cancer, and prostate cancer. However, LDR and pulsed-dose-rate interstitial brachytherapies have been mainstays for head and neck cancer. HDR brachytherapy has not become widely used in the radiotherapy community for treating head and neck cancer because of lack of experience and biological concerns. On the other hand, because HDR brachytherapy is less time-consuming, treatment can occasionally be administered on an outpatient basis. For the convenience and safety of patients and medical staff, HDR brachytherapy should be explored. To enhance the role of this therapy in treatment of head and neck lesions, we have reviewed its outcomes with oral cancer, including Phase I/II to Phase III studies, evaluating this technique in terms of safety and efficacy. In particular, our studies have shown that superficial tumors can be treated using a non-invasive mold technique on an outpatient basis without adverse reactions. The next generation of image-guided brachytherapy using HDR has been discussed. In conclusion, although concrete evidence is yet to be produced with a sophisticated study in a reproducible manner, HDR brachytherapy remains an important option for treatment of oral cancer. (author)

  2. An automated optimization tool for high-dose-rate (HDR) prostate brachytherapy with divergent needle pattern

    Science.gov (United States)

    Borot de Battisti, M.; Maenhout, M.; de Senneville, B. Denis; Hautvast, G.; Binnekamp, D.; Lagendijk, J. J. W.; van Vulpen, M.; Moerland, M. A.

    2015-10-01

    Focal high-dose-rate (HDR) for prostate cancer has gained increasing interest as an alternative to whole gland therapy as it may contribute to the reduction of treatment related toxicity. For focal treatment, optimal needle guidance and placement is warranted. This can be achieved under MR guidance. However, MR-guided needle placement is currently not possible due to space restrictions in the closed MR bore. To overcome this problem, a MR-compatible, single-divergent needle-implant robotic device is under development at the University Medical Centre, Utrecht: placed between the legs of the patient inside the MR bore, this robot will tap the needle in a divergent pattern from a single rotation point into the tissue. This rotation point is just beneath the perineal skin to have access to the focal prostate tumor lesion. Currently, there is no treatment planning system commercially available which allows optimization of the dose distribution with such needle arrangement. The aim of this work is to develop an automatic inverse dose planning optimization tool for focal HDR prostate brachytherapy with needle insertions in a divergent configuration. A complete optimizer workflow is proposed which includes the determination of (1) the position of the center of rotation, (2) the needle angulations and (3) the dwell times. Unlike most currently used optimizers, no prior selection or adjustment of input parameters such as minimum or maximum dose or weight coefficients for treatment region and organs at risk is required. To test this optimizer, a planning study was performed on ten patients (treatment volumes ranged from 8.5 cm3to 23.3 cm3) by using 2-14 needle insertions. The total computation time of the optimizer workflow was below 20 min and a clinically acceptable plan was reached on average using only four needle insertions.

  3. Dosimetric and radiobiological comparison of volumetric modulated arc therapy, high-dose rate brachytherapy, and low-dose rate permanent seeds implant for localized prostate cancer

    Energy Technology Data Exchange (ETDEWEB)

    Yang, Ruijie, E-mail: ruijyang@yahoo.com; Zhao, Nan; Liao, Anyan; Wang, Hao; Qu, Ang

    2016-10-01

    To investigate the dosimetric and radiobiological differences among volumetric modulated arc therapy (VMAT), high-dose rate (HDR) brachytherapy, and low-dose rate (LDR) permanent seeds implant for localized prostate cancer. A total of 10 patients with localized prostate cancer were selected for this study. VMAT, HDR brachytherapy, and LDR permanent seeds implant plans were created for each patient. For VMAT, planning target volume (PTV) was defined as the clinical target volume plus a margin of 5 mm. Rectum, bladder, urethra, and femoral heads were considered as organs at risk. A 78 Gy in 39 fractions were prescribed for PTV. For HDR and LDR plans, the dose prescription was D{sub 90} of 34 Gy in 8.5 Gy per fraction, and 145 Gy to clinical target volume, respectively. The dose and dose volume parameters were evaluated for target, organs at risk, and normal tissue. Physical dose was converted to dose based on 2-Gy fractions (equivalent dose in 2 Gy per fraction, EQD{sub 2}) for comparison of 3 techniques. HDR and LDR significantly reduced the dose to rectum and bladder compared with VMAT. The D{sub mean} (EQD{sub 2}) of rectum decreased 22.36 Gy in HDR and 17.01 Gy in LDR from 30.24 Gy in VMAT, respectively. The D{sub mean} (EQD{sub 2}) of bladder decreased 6.91 Gy in HDR and 2.53 Gy in LDR from 13.46 Gy in VMAT. For the femoral heads and normal tissue, the mean doses were also significantly reduced in both HDR and LDR compared with VMAT. For the urethra, the mean dose (EQD{sub 2}) was 80.26, 70.23, and 104.91 Gy in VMAT, HDR, and LDR brachytherapy, respectively. For localized prostate cancer, both HDR and LDR brachytherapy were clearly superior in the sparing of rectum, bladder, femoral heads, and normal tissue compared with VMAT. HDR provided the advantage in sparing of urethra compared with VMAT and LDR.

  4. Dosimetric and radiobiological comparison of volumetric modulated arc therapy, high-dose rate brachytherapy, and low-dose rate permanent seeds implant for localized prostate cancer

    International Nuclear Information System (INIS)

    Yang, Ruijie; Zhao, Nan; Liao, Anyan; Wang, Hao; Qu, Ang

    2016-01-01

    To investigate the dosimetric and radiobiological differences among volumetric modulated arc therapy (VMAT), high-dose rate (HDR) brachytherapy, and low-dose rate (LDR) permanent seeds implant for localized prostate cancer. A total of 10 patients with localized prostate cancer were selected for this study. VMAT, HDR brachytherapy, and LDR permanent seeds implant plans were created for each patient. For VMAT, planning target volume (PTV) was defined as the clinical target volume plus a margin of 5 mm. Rectum, bladder, urethra, and femoral heads were considered as organs at risk. A 78 Gy in 39 fractions were prescribed for PTV. For HDR and LDR plans, the dose prescription was D 90 of 34 Gy in 8.5 Gy per fraction, and 145 Gy to clinical target volume, respectively. The dose and dose volume parameters were evaluated for target, organs at risk, and normal tissue. Physical dose was converted to dose based on 2-Gy fractions (equivalent dose in 2 Gy per fraction, EQD 2 ) for comparison of 3 techniques. HDR and LDR significantly reduced the dose to rectum and bladder compared with VMAT. The D mean (EQD 2 ) of rectum decreased 22.36 Gy in HDR and 17.01 Gy in LDR from 30.24 Gy in VMAT, respectively. The D mean (EQD 2 ) of bladder decreased 6.91 Gy in HDR and 2.53 Gy in LDR from 13.46 Gy in VMAT. For the femoral heads and normal tissue, the mean doses were also significantly reduced in both HDR and LDR compared with VMAT. For the urethra, the mean dose (EQD 2 ) was 80.26, 70.23, and 104.91 Gy in VMAT, HDR, and LDR brachytherapy, respectively. For localized prostate cancer, both HDR and LDR brachytherapy were clearly superior in the sparing of rectum, bladder, femoral heads, and normal tissue compared with VMAT. HDR provided the advantage in sparing of urethra compared with VMAT and LDR.

  5. Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry

    International Nuclear Information System (INIS)

    Casey, Kevin E.; Kry, Stephen F.; Howell, Rebecca M.; Followill, David; Alvarez, Paola; Lawyer, Ann

    2013-01-01

    Purpose: The aim of this work was to create a mailable phantom with measurement accuracy suitable for Radiological Physics Center (RPC) audits of high dose-rate (HDR) brachytherapy sources at institutions participating in National Cancer Institute-funded cooperative clinical trials. Optically stimulated luminescence dosimeters (OSLDs) were chosen as the dosimeter to be used with the phantom.Methods: The authors designed and built an 8 × 8 × 10 cm 3 prototype phantom that had two slots capable of holding Al 2 O 3 :C OSLDs (nanoDots; Landauer, Glenwood, IL) and a single channel capable of accepting all 192 Ir HDR brachytherapy sources in current clinical use in the United States. The authors irradiated the phantom with Nucletron and Varian 192 Ir HDR sources in order to determine correction factors for linearity with dose and the combined effects of irradiation energy and phantom characteristics. The phantom was then sent to eight institutions which volunteered to perform trial remote audits.Results: The linearity correction factor was k L = (−9.43 × 10 −5 × dose) + 1.009, where dose is in cGy, which differed from that determined by the RPC for the same batch of dosimeters using 60 Co irradiation. Separate block correction factors were determined for current versions of both Nucletron and Varian 192 Ir HDR sources and these vendor-specific correction factors differed by almost 2.6%. For the Nucletron source, the correction factor was 1.026 [95% confidence interval (CI) = 1.023–1.028], and for the Varian source, it was 1.000 (95% CI = 0.995–1.005). Variations in lateral source positioning up to 0.8 mm and distal/proximal source positioning up to 10 mm had minimal effect on dose measurement accuracy. The overall dose measurement uncertainty of the system was estimated to be 2.4% and 2.5% for the Nucletron and Varian sources, respectively (95% CI). This uncertainty was sufficient to establish a ±5% acceptance criterion for source strength audits under a

  6. Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Casey, Kevin E.; Kry, Stephen F.; Howell, Rebecca M.; Followill, David [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030 and The University of Texas Graduate School of Biomedical Sciences at Houston, Houston, Texas 77030 (United States); Alvarez, Paola; Lawyer, Ann [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030 (United States)

    2013-11-15

    Purpose: The aim of this work was to create a mailable phantom with measurement accuracy suitable for Radiological Physics Center (RPC) audits of high dose-rate (HDR) brachytherapy sources at institutions participating in National Cancer Institute-funded cooperative clinical trials. Optically stimulated luminescence dosimeters (OSLDs) were chosen as the dosimeter to be used with the phantom.Methods: The authors designed and built an 8 × 8 × 10 cm{sup 3} prototype phantom that had two slots capable of holding Al{sub 2}O{sub 3}:C OSLDs (nanoDots; Landauer, Glenwood, IL) and a single channel capable of accepting all {sup 192}Ir HDR brachytherapy sources in current clinical use in the United States. The authors irradiated the phantom with Nucletron and Varian {sup 192}Ir HDR sources in order to determine correction factors for linearity with dose and the combined effects of irradiation energy and phantom characteristics. The phantom was then sent to eight institutions which volunteered to perform trial remote audits.Results: The linearity correction factor was k{sub L}= (−9.43 × 10{sup −5}× dose) + 1.009, where dose is in cGy, which differed from that determined by the RPC for the same batch of dosimeters using {sup 60}Co irradiation. Separate block correction factors were determined for current versions of both Nucletron and Varian {sup 192}Ir HDR sources and these vendor-specific correction factors differed by almost 2.6%. For the Nucletron source, the correction factor was 1.026 [95% confidence interval (CI) = 1.023–1.028], and for the Varian source, it was 1.000 (95% CI = 0.995–1.005). Variations in lateral source positioning up to 0.8 mm and distal/proximal source positioning up to 10 mm had minimal effect on dose measurement accuracy. The overall dose measurement uncertainty of the system was estimated to be 2.4% and 2.5% for the Nucletron and Varian sources, respectively (95% CI). This uncertainty was sufficient to establish a ±5% acceptance

  7. Variability of Marker-Based Rectal Dose Evaluation in HDR Cervical Brachytherapy

    International Nuclear Information System (INIS)

    Wang Zhou; Jaggernauth, Wainwright; Malhotra, Harish K.; Podgorsak, Matthew B.

    2010-01-01

    In film-based intracavitary brachytherapy for cervical cancer, position of the rectal markers may not accurately represent the anterior rectal wall. This study was aimed at analyzing the variability of rectal dose estimation as a result of interfractional variation of marker placement. A cohort of five patients treated with multiple-fraction tandem and ovoid high-dose-rate (HDR) brachytherapy was studied. The cervical os point and the orientation of the applicators were matched among all fractional plans for each patient. Rectal points obtained from all fractions were then input into each clinical treated plan. New fractional rectal doses were obtained and a new cumulative rectal dose for each patient was calculated. The maximum interfractional variation of distances between rectal dose points and the closest source positions was 1.1 cm. The corresponding maximum variability of fractional rectal dose was 65.5%. The percentage difference in cumulative rectal dose estimation for each patient was 5.4%, 19.6%, 34.6%, 23.4%, and 13.9%, respectively. In conclusion, care should be taken when using rectal markers as reference points for estimating rectal dose in HDR cervical brachytherapy. The best estimate of true rectal dose for each fraction should be determined by the most anterior point among all fractions.

  8. Use of CT or MR dosimetry in high dose rate (HDR) brachytherapy for prostate cancer

    International Nuclear Information System (INIS)

    Liu, C.; Das, R.; See, A.; Duchesne, G.M.; Van Dyk, S.; Tai, K.H.

    2003-01-01

    Brachytherapy (BT) has, in recent years, become a well-utilised treatment option for prostate cancer. Tumour control probability relies on accurate dosimetry, which in turn relies on the accurate definition of the prostate gland. In external beam radiotherapy and BT, MRI has been shown to be a superior imaging modality when delineating the prostate gland especially at the apex. To date, data on MRI planning in prostate BT has focussed mainly on permanent interstitial implants. No data currently exists comparing MRI vs CT planning in HDR BT and its subsequent impact on prostate dosimetry. To determine the effects of MRI vs CT in HDR BT with respect to prostatic volumes and normal tissue doses, with the evaluations made using dose-volume histograms (DVH). Dosimetry parameters derived using CT and MRI (T2 weighted) scans of 11 patients who had received TRUS guided implants for HDR BT, were compared using the PlatoTM computer planning system. Treatment plans were generated on volumes marked by the same radiation oncologist for each patient. Comparison was made of the treatment plans (dosimetry) between: 1. CT generated plans; 2. CT generated plans assessed using MRI marked volumes and 3. MRI generated plans. We confirm the previously reported results that CT scans can overestimate prostatic volumes compared with MRI. Variations were noted in CT and MRI based plans that may allow improved sparing of the rectum and urethra when using MRI planning. The main disadvantages of using MRI scans are access to facilities as well as identifying a dummy source to adequately define the tips of our catheters. It is feasible to utilise MRI scans for HDR BT planning. The clearer definition of anatomical structures has added advantages when contouring the prostate

  9. A comparison of HDR brachytherapy and IMRT techniques for dose escalation in prostate cancer: A radiobiological modeling study

    Energy Technology Data Exchange (ETDEWEB)

    Fatyga, M.; Williamson, J. F.; Dogan, N.; Todor, D.; Siebers, J. V.; George, R.; Barani, I.; Hagan, M. [Department of Radiation Oncology, Virginia Commonwealth University Medical Center, 401 College Street, Richmond, Virginia 23298 (United States)

    2009-09-15

    A course of one to three large fractions of high dose rate (HDR) interstitial brachytherapy is an attractive alternative to intensity modulated radiation therapy (IMRT) for delivering boost doses to the prostate in combination with additional external beam irradiation for intermediate risk disease. The purpose of this work is to quantitatively compare single-fraction HDR boosts to biologically equivalent fractionated IMRT boosts, assuming idealized image guided delivery (igIMRT) and conventional delivery (cIMRT). For nine prostate patients, both seven-field IMRT and HDR boosts were planned. The linear-quadratic model was used to compute biologically equivalent dose prescriptions. The cIMRT plan was evaluated as a static plan and with simulated random and setup errors. The authors conclude that HDR delivery produces a therapeutic ratio which is significantly better than the conventional IMRT and comparable to or better than the igIMRT delivery. For the HDR, the rectal gBEUD analysis is strongly influenced by high dose DVH tails. A saturation BED, beyond which no further injury can occur, must be assumed. Modeling of organ motion uncertainties yields mean outcomes similar to static plan outcomes.

  10. High dose rate versus low dose rate brachytherapy for oral cancer--a meta-analysis of clinical trials.

    Directory of Open Access Journals (Sweden)

    Zhenxing Liu

    Full Text Available To compare the efficacy and safety of high dose rate (HDR and low dose rate (LDR brachytherapy in treating early-stage oral cancer.A systematic search of MEDLINE, EMBASE and Cochrane Library databases, restricted to English language up to June 1, 2012, was performed to identify potentially relevant studies.Only randomized controlled trials (RCT and controlled trials that compared HDR to LDR brachytherapy in treatment of early-stage oral cancer (stages I, II and III were of interest.Two investigators independently extracted data from retrieved studies and controversies were solved by discussion. Meta-analysis was performed using RevMan 5.1. One RCT and five controlled trials (607 patients: 447 for LDR and 160 for HDR met the inclusion criteria. The odds ratio showed no statistically significant difference between LDR group and HDR group in terms of local recurrence (OR = 1.12, CI 95% 0.62-2.01, overall mortality (OR = 1.01, CI 95% 0.61-1.66 and Grade 3/4 complications (OR = 0.86, CI 95% 0.52-1.42.This meta-analysis indicated that HDR brachytherapy was a comparable alternative to LDR brachytherapy in treatment of oral cancer. HDR brachytherapy might become a routine choice for early-stage oral cancer in the future.

  11. High-dose-rate versus low-dose-rate brachytherapy in the treatment of cervical cancer: analysis of tumor recurrence - the University of Wisconsin experience

    International Nuclear Information System (INIS)

    Petereit, Daniel G.; Sarkaria, Jann N.; Potter, David M.; Schink, Julian C.

    1999-01-01

    Purpose: To retrospectively compare the clinical outcome for cervical cancer patients treated with high-dose-rate (HDR) vs. low-dose-rate (LDR) brachytherapy. Methods and Materials: One hundred ninety-one LDR patients were treated from 1977 to 1988 and compared to 173 HDR patients treated from 1989 to 1996. Patients of similar stage and tumor volumes were treated with identical external beam fractionation schedules. Brachytherapy was given in either 1 or 2 LDR implants for the earlier patient cohort, and 5 HDR implants for the latter cohort. For both patient groups, Point A received a minimum total dose of 80 Gy. The linear-quadratic formula was used to calculate the LDR dose-equivalent contribution to Point A for the HDR treatments. The primary endpoints assessed were survival, pelvic control, relapse-free survival, and distant metastases. Endpoints were estimated using the Kaplan-Meier method. Comparisons between treatment groups were performed using the log-rank test and Cox proportional hazards models. Results: The median follow-up was 65 months (2 to 208 months) in the LDR group and 22 months (1 to 85 months) in the HDR group. For all stages combined there was no difference in survival, pelvic control, relapse-free survival, or distant metastases between LDR and HDR patients. For Stage IB and II HDR patients, the pelvic control rates were 85% and 80% with survival rates of 86% and 65% at 3 years, respectively. In the LDR group, Stage IB and II patients had 91% and 78% pelvic control rates, with 82% and 58% survival rates at 3 years, respectively. No difference was seen in survival or pelvic control for bulky Stage I and II patients combined (> 5 cm). Pelvic control at 3 years was 44% (HDR) versus 75% (LDR) for Stage IIIB patients (p = 0.002). This difference in pelvic control was associated with a lower survival rate in the Stage IIIB HDR versus LDR population (33% versus 58%, p = 0.004). The only major difference, with regard to patient characteristics

  12. Ir-192 HDR transit dose and radial dose function determination using alanine/EPR dosimetry

    International Nuclear Information System (INIS)

    Calcina, Carmen S Guzman; Almeida, Adelaide de; Rocha, Jose R Oliveira; Abrego, Felipe Chen; Baffa, Oswaldo

    2005-01-01

    Source positioning close to the tumour in high dose rate (HDR) brachytherapy is not instantaneous. An increment of dose will be delivered during the movement of the source in the trajectory to its static position. This increment is the transit dose, often not taken into account in brachytherapeutic treatment planning. The transit dose depends on the prescribed dose, number of treatment fractions, velocity and activity of the source. Combining all these factors, the transit dose can be 5% higher than the prescribed absorbed dose value (Sang-Hyun and Muller-Runkel, 1994 Phys. Med. Biol. 39 1181-8, Nath et al 1995 Med. Phys. 22 209-34). However, it cannot exceed this percentage (Nath et al 1995). In this work, we use the alanine-EPR (electron paramagnetic resonance) dosimetric system using analysis of the first derivative of the signal. The transit dose was evaluated for an HDR system and is consistent with that already presented for TLD dosimeters (Bastin et al 1993 Int. J. Radiat. Oncol. Biol. Phys. 26 695-702). Also using the same dosimetric system, the radial dose function, used to evaluate the geometric dose degradation around the source, was determined and its behaviour agrees better with those obtained by Monte Carlo simulations (Nath et al 1995, Williamson and Nath 1991 Med. Phys. 18 434-48, Ballester et al 1997 Med. Phys. 24 1221-8, Ballester et al 2001 Phys. Med. Biol. 46 N79-90) than with TLD measurements (Nath et al 1990 Med. Phys. 17 1032-40)

  13. High dose rate intraoperative radiation therapy (HDR-IORT) as part of the management strategy for locally advanced primary and recurrent rectal cancer

    International Nuclear Information System (INIS)

    Harrison, Louis B.; Minsky, Bruce D.; Enker, Warren E.; Mychalczak, Borys; Guillem, Jose; Paty, Philip B.; Anderson, Lowell; White, Carol; Cohen, Alfred M.

    1998-01-01

    Purpose: Primary unresectable and locally advanced recurrent rectal cancer presents a significant clinical challenge. Local failure rates are high in both situations. Under such circumstances, there is a significant need to safely deliver tumoricidal doses of radiation in an attempt to improve local control. For this reason, we have incorporated a new approach utilizing high dose rate intraoperative radiation therapy (HDR-IORT). Methods and Materials: Between 11/92-12/96, a total of 112 patients were explored, of which 68 patients were treated with HDR-IORT, and 66 are evaluable. The majority of the 44 patients were excluded for unresectable disease or for distant metastases which eluded preoperative imaging. There were 22 patients with primary unresectable disease, and 46 patients who presented with recurrent disease. The histology was adenocarcinoma in 64 patients, and squamous cell carcinoma in four patients. In general, the patients with primary unresectable disease received preoperative chemotherapy with 5-fluorouracil (5-FU) and leucovorin, and external beam irradiation to 4500-5040 cGy, followed by surgical resection and HDR-IORT (1000-2000 cGy). In general , the patients with recurrent disease were treated with surgical resection and HDR-IORT (1000-2000 cGy) alone. All surgical procedures were done in a dedicated operating room in the brachytherapy suite, so that HDR-IORT could be delivered using the Harrison-Anderson-Mick (HAM) applicator. The median follow-up is 17.5 months (1-48 mo). Results: In primary cases, the actuarial 2-year local control is 81%. For patients with negative margins, the local control was 92% vs. 38% for those with positive margins (p = 0.002). The 2-year actuarial disease-free survival was 69%; 77% for patients with negative margins vs. 38% for patients with positive margins (p = 0.03). For patients with recurrent disease, the 2-year actuarial local control rate was 63%. For patients with negative margins, it was 82%, while it was

  14. Time-driven activity-based costing of low-dose-rate and high-dose-rate brachytherapy for low-risk prostate cancer.

    Science.gov (United States)

    Ilg, Annette M; Laviana, Aaron A; Kamrava, Mitchell; Veruttipong, Darlene; Steinberg, Michael; Park, Sang-June; Burke, Michael A; Niedzwiecki, Douglas; Kupelian, Patrick A; Saigal, Christopher

    Cost estimates through traditional hospital accounting systems are often arbitrary and ambiguous. We used time-driven activity-based costing (TDABC) to determine the true cost of low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy for prostate cancer and demonstrate opportunities for cost containment at an academic referral center. We implemented TDABC for patients treated with I-125, preplanned LDR and computed tomography based HDR brachytherapy with two implants from initial consultation through 12-month followup. We constructed detailed process maps for provision of both HDR and LDR. Personnel, space, equipment, and material costs of each step were identified and used to derive capacity cost rates, defined as price per minute. Each capacity cost rate was then multiplied by the relevant process time and products were summed to determine total cost of care. The calculated cost to deliver HDR was greater than LDR by $2,668.86 ($9,538 vs. $6,869). The first and second HDR treatment day cost $3,999.67 and $3,955.67, whereas LDR was delivered on one treatment day and cost $3,887.55. The greatest overall cost driver for both LDR and HDR was personnel at 65.6% ($4,506.82) and 67.0% ($6,387.27) of the total cost. After personnel costs, disposable materials contributed the second most for LDR ($1,920.66, 28.0%) and for HDR ($2,295.94, 24.0%). With TDABC, the true costs to deliver LDR and HDR from the health system perspective were derived. Analysis by physicians and hospital administrators regarding the cost of care afforded redesign opportunities including delivering HDR as one implant. Our work underscores the need to assess clinical outcomes to understand the true difference in value between these modalities. Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  15. 192Ir high dose rate (HDR) interstitial brain implant: optimisation

    International Nuclear Information System (INIS)

    Tyagi, Anuj; Singh, Dinesh; Chitra, S.; Gupta, J.P.

    2001-01-01

    The new modality of stepping source dosimetry system (SSDs) illustrates a remarkable improvement in attaining the uniform and homogeneous dose distribution within the target volume. The technique enables the physicist to correct for a certain amount of misplacement or curvature of implant geometry. The short course of brachytherapy provides good palliation in terms of functional improvements with low and acceptable toxicity in high-grade glioma. With continual refinements of the technique, brachytherapy performed by a skilled brachytherapy team offers an opportunity to improve patient survival and quality of life. Since 1997, micro selectron HDR 192 Ir treatments are done including gynecological, oesophageal, breast, surface mould, soft tissue sarcoma (STS) and brain in our hospital. In this paper, procedure of interstitial brain implant in glioma as implant technique, simulation and treatment planning will be discussed

  16. Estimation of the transit dose component in high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Garcia Romero, A.; Millan Cebrian, E.; Lozano Flores, F.J.; Lope Lope, R.; Canellas Anoz, M.

    2001-01-01

    Current high dose rate brachytherapy (HDR) treatment planning systems usually calculate dose only from source stopping positions (stationary component), but fails to account for the administered dose when the source is moving (dynamic component or transit dose). Numerical values of this transit dose depends upon the source velocity, implant geometry, source activity and prescribed dose. In some HDR treatments using particular geometry the transit dose cannot be ignored because it increases the dose at the prescriptions points and also could increase potential late tissue complications as predicted by the linear quadratic model. International protocols recommend to verify this parameter. The aim of this paper has been to establish a procedure for the transit dose calculation for the Gammamed 12i equipment at the RT Department in the Clinical University Hospital (Zaragoza-Spain). A numeric algorithm was implemented based on a dynamic point approximation for the moving HDR source and the calculated results for the entrance-exit transit dose was compared with TLD measurements made in some discrete points. (author) [es

  17. On-line conformal HDR dose escalation trial in prostate cancer

    International Nuclear Information System (INIS)

    Martinez, Alvaro; Stromberg, Jannifer; Edmundson, Gregory; Gustafson, Gary; Vicini, Frank; Brabbins, Donald

    1996-01-01

    Purpose: To improve treatment results on prostatic adenocarcinoma, we began the first prospective Phase I/II dose-escalating clinical trial of conformal brachytherapy (CB) and concurrent external beam irradiation. Methods and Materials: Fifty-four patients with T2b-T3c prostatic adenocarcinoma received 172 transperineal conformal high-dose rate (HDR) boost implants. All patients received concomitant external beam pelvic irradiation. Dose escalation of the three HDR fractions were: 5.5 Gy (18 patients), 6 Gy (15 patients), and 6.5 Gy (21 patients). The urethra, anterior rectal wall, and prostate boundaries were identified individually and outlined at 5 mm intervals from the base to the apex of the gland. The CB using real-time ultrasound guidance with interactive online isodose distributions was performed on an outpatient basis. As needles were placed into the prostate, corrections for prostate displacement were recorded and the isodose distributions were recalculated to represent the new relationship between the needles, prostate, and normal structures. Results: Craniocaudal motion of the gland ranged from 0.5-2.0 cm (mean=1.0 cm), whereas lateral displacement was 0.1-0.4 cm. With the interactive online planning system, organ motion was immediately detected, accounted for, and corrected prior to each HDR treatment. The rectal dose has ranged from 45 to 87%, and the urethral dose from 97 to 112% of the prostate dose. Negative prostatic biopsies at 18 months were seen in (30(32)) patients. Biochemical (PSA <1.5 ng/ml) control at 36 months is is 89%. It is significant that operator dependence has been completely removed because the interactive online planning system uniformly guides the physicians. Conclusions: With ultrasound guidance and the interactive online dosimetry system, organ motion is insignificant because it can be corrected during the procedure. Common pitfalls of brachytherapy, including operator dependence and difficulty with reproducibility, have been

  18. Genotoxic effects of high dose rate X-ray and low dose rate gamma radiation in ApcMin/+ mice.

    Science.gov (United States)

    Graupner, Anne; Eide, Dag M; Brede, Dag A; Ellender, Michele; Lindbo Hansen, Elisabeth; Oughton, Deborah H; Bouffler, Simon D; Brunborg, Gunnar; Olsen, Ann Karin

    2017-10-01

    Risk estimates for radiation-induced cancer in humans are based on epidemiological data largely drawn from the Japanese atomic bomb survivor studies, which received an acute high dose rate (HDR) ionising radiation. Limited knowledge exists about the effects of chronic low dose rate (LDR) exposure, particularly with respect to the application of the dose and dose rate effectiveness factor. As part of a study to investigate the development of colon cancer following chronic LDR vs. acute HDR radiation, this study presents the results of genotoxic effects in blood of exposed mice. CBAB6 F1 Apc +/+ (wild type) and Apc Min/+ mice were chronically exposed to estimated whole body absorbed doses of 1.7 or 3.2 Gy 60 Co-γ-rays at a LDR (2.2 mGy h -1 ) or acutely exposed to 2.6 Gy HDR X-rays (1.3 Gy min -1 ). Genotoxic endpoints assessed in blood included chromosomal damage (flow cytometry based micronuclei (MN) assay), mutation analyses (Pig-a gene mutation assay), and levels of DNA lesions (Comet assay, single-strand breaks (ssb), alkali labile sites (als), oxidized DNA bases). Ionising radiation (ca. 3 Gy) induced genotoxic effects dependent on the dose rate. Chromosomal aberrations (MN assay) increased 3- and 10-fold after chronic LDR and acute HDR, respectively. Phenotypic mutation frequencies as well as DNA lesions (ssb/als) were modulated after acute HDR but not after chronic LDR. The Apc Min/+ genotype did not influence the outcome in any of the investigated endpoints. The results herein will add to the scant data available on genotoxic effects following chronic LDR of ionising radiation. Environ. Mol. Mutagen. 58:560-569, 2017. © 2017 The Authors Environmental and Molecular Mutagenesis published by Wiley Periodicals, Inc. on behalf of Environmental Mutagen Society. © 2017 The Authors Environmental and Molecular Mutagenesis published by Wiley Periodicals, Inc. on behalf of Environmental Mutagen Society.

  19. Late complications after high-dose-rate interstitial brachytherapy for tongue cancer

    International Nuclear Information System (INIS)

    Shimizutani, Kimishige; Inoue, Takehiro; Inoue, Toshihiko; Yoshioka, Yasuo; Teshima, Teruki; Kakimoto, Naoya; Murakami, Shumei; Furukawa, Souhei; Fuchihata, Hajime

    2005-01-01

    The objectives of this study was to analyze the treatment results and late complications of high-dose-rate (HDR) interstitial brachytherapy (ISBT) for early (T1N0, T2N0) mobile tongue cancer using the microSelectron-HDR. From January 1993 through April 2001, a total of 72 patients with early squamous cell carcinomas of the mobile tongue were treated with microSelectron-HDR interstitial brachytherapy at the Department of Radiology, Osaka University Hospital. Of the patients, 18% were treated with a combination of prior external radiation and HDR-ISBT, and 82% were treated with HDR-ISBT alone. For HDR-ISBT alone, all cases were treated with a total dose of 54 Gy/9 fractions every 5 days or 60 Gy/10 fractions every 8 days. In combined therapy with an external dose of 30 to 40 Gy, HDR-ISBT was given at a total dose of 42-50 Gy. The Brinkman and alcohol indexes were used to analyze the incidence of late complications after HDR-ISBT. The 2- and 5-year local control rates were 85% and 82%, respectively. Fifteen of 72 patients (21%) treated with HDR-ISBT had late complications. Ten of 15 patients (67%) with late complications had a Brinkman index exceeding 600. HDR-ISBT is useful and easily applied under local anesthesia to early or superficial lesions of the mobile tongue. However, we found an increase in late complications, such as soft-tissue ulcers and bone exposure, after irradiation of tongue cancer with 60 Gy HDR-ISBT in patients with a Brinkman index greater than 600. (author)

  20. Radiobiological considerations in gynaecological HDR and LDR brachytherapy

    International Nuclear Information System (INIS)

    Bauer, M.; Schulz-Wendtland, R.

    1989-01-01

    In brachytherapy the advantages of high dose rate over low dose rate afterloading therapy were obvious. Out-patient treatment becomes possible, the position of the sources is reproducible and can be observed during the treatment and the patients have to be immobilised for only a short time, giving less psychological stress and a decreased risk of thrombosis and embolism. When changing from LDR to HDR afterloading therapy we are not yet able to evaluate its biological impact. Radiobiological considerations and our experimental data, however, give us the following clinical consequences by using HDR brachytherapy: There is a need for about 15 fractions or more and each increase in dose rate requires higher fractioning. Due to the steep dose rate decline and the inhomogeneous dose distribution, multiple equivalence factors are necessary when fractioning is not sufficiently high. Correction factors to reduce the dose close to the source are low, with increasing distance from the source they increase. If HDR radiation therapy is used, the percutaneous dose in the pelvic wall region should be reduced. The reduction of the dose in HDR brachytherapy is a compromise to limit the side effects caused by the radiation. The drawback is a small therapeutic range and reduced therapeutic effectivity at the tumour. (orig.) [de

  1. HIGH-DOSE RATE BRACHYTHERAPY IN CARCINOMA CERVIX STAGE IIIB

    Directory of Open Access Journals (Sweden)

    Sathya Maruthavanan

    2016-07-01

    Full Text Available INTRODUCTION Radiotherapy is the standard treatment in locally advanced (IIB-IVA and early inoperable cases. The current standard of practice with curable intent is concurrent chemoradiation in which intracavitary brachytherapy is an integral component of radiotherapy. This study aims at assessing the efficacy of HDR ICBT (High-dose rate intracavitary brachytherapy in terms local response, normal tissue reactions, and feasibility. METHODS AND MATERIALS A total of 20 patients of stage IIIB cancer of the uterine cervix were enrolled in the study and were planned to receive concurrent chemotherapy weekly along with EBRT (external beam radiotherapy to a dose of 50 Gy/25 Fr. Suitability for ICBT was assessed at 40 Gy/20 Fr. 6/20 patients were suitable at 40 Gy and received HDR ICBT with a dose of 5.5 Gy to point A in 4 sessions (5.5 Gy/4 Fr. The remaining 14/20 patients completed 50 Gy and received HDR ICBT with a dose of 6 Gy to point A in 3 sessions (6 Gy/3 Fr. RESULTS A total of 66 intracavitary applications were done and only one application required dose modification due to high bladder dose, the pelvic control rate was 85% (17/20. 10% (2/20 had stable disease and 5% (1/20 had progressive disease at one year of follow up. When toxicity was considered only 15% developed grade I and grade II rectal complications. Patient compliance and acceptability was 100%. Patients were very comfortable with the short treatment time as compared with patients on LDR ICBT (low-dose rate intracavitary brachytherapy treatment interviewed during the same period. CONCLUSION This study proves that HDR brachytherapy is efficacious and feasible in carcinoma of cervix stage IIIB. It also proves that good dose distribution can be achieved with HDR intracavitary facility by the use of dose optimization. The short treatment time in HDR ICBT makes it possible to maintain this optimised dose distribution throughout the treatment providing a gain in the therapeutic ratio and

  2. Evaluation of radiation doses on critical organs in the treatment of cancer of the cervix using HDR-brachytherapy

    International Nuclear Information System (INIS)

    Soares, Taciana; Jansem, Teresa

    2000-01-01

    High dose-rate (HDR) brachytherapy is one type of treatment of the cervix carcinoma. During the planning for this therapy, especial attention is given to proximal normal organs such as bladder and rectum. In fact, due to their radiosensibility and localization, bladder and rectum are considered as critical organs. In this work we have studied the influence of the positioning of patient legs in the dose delivered to these critical organs in the treatment of cancer of the cervix using HDR-brachytherapy. (author)

  3. Brachytherapy for cervix cancer: low-dose rate or high-dose rate brachytherapy – a meta-analysis of clinical trials

    Directory of Open Access Journals (Sweden)

    Stefano Eduardo J

    2009-04-01

    Full Text Available Abstract Background The literature supporting high-dose rate brachytherapy (HDR in the treatment of cervical carcinoma derives primarily from retrospective series. However, controversy still persists regarding the efficacy and safety of HDR brachytherapy compared to low-dose rate (LDR brachytherapy, in particular, due to inadequate tumor coverage for stage III patients. Whether LDR or HDR brachytherapy produces better results for these patients in terms of survival rate, local control rate and the treatment complications remain controversial. Methods A meta-analysis of RCT was performed comparing LDR to HDR brachytherapy for cervix cancer treated for radiotherapy alone. The MEDLINE, EMBASE, CANCERLIT and Cochrane Library databases, as well as abstracts published in the annual proceedings were systematically searched. We assessed methodological quality for each outcome by grading the quality of evidence using the Grading of Recommendations Assessment, Development and Evaluation (GRADE methodology. We used "recommend" for strong recommendations, and "suggest" for weak recommendations. Results Pooled results from five randomized trials (2,065 patients of HDR brachytherapy in cervix cancer showed no significant increase of mortality (p = 0.52, local recurrence (p = 0.68, or late complications (rectal; p = 0.7, bladder; p = 0.95 or small intestine; p = 0.06 rates as compared to LDR brachytherapy. In the subgroup analysis no difference was observed for overall mortality and local recurrence in patients with clinical stages I, II and III. The quality of evidence was low for mortality and local recurrence in patients with clinical stage I, and moderate for other clinical stages. Conclusion Our meta-analysis shows that there are no differences between HDR and LDR for overall survival, local recurrence and late complications for clinical stages I, II and III. By means of the GRADE system, we recommend the use of HDR for all clinical stages of cervix

  4. Brachytherapy for cervix cancer: low-dose rate or high-dose rate brachytherapy – a meta-analysis of clinical trials

    Science.gov (United States)

    Viani, Gustavo A; Manta, Gustavo B; Stefano, Eduardo J; de Fendi, Ligia I

    2009-01-01

    Background The literature supporting high-dose rate brachytherapy (HDR) in the treatment of cervical carcinoma derives primarily from retrospective series. However, controversy still persists regarding the efficacy and safety of HDR brachytherapy compared to low-dose rate (LDR) brachytherapy, in particular, due to inadequate tumor coverage for stage III patients. Whether LDR or HDR brachytherapy produces better results for these patients in terms of survival rate, local control rate and the treatment complications remain controversial. Methods A meta-analysis of RCT was performed comparing LDR to HDR brachytherapy for cervix cancer treated for radiotherapy alone. The MEDLINE, EMBASE, CANCERLIT and Cochrane Library databases, as well as abstracts published in the annual proceedings were systematically searched. We assessed methodological quality for each outcome by grading the quality of evidence using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) methodology. We used "recommend" for strong recommendations, and "suggest" for weak recommendations. Results Pooled results from five randomized trials (2,065 patients) of HDR brachytherapy in cervix cancer showed no significant increase of mortality (p = 0.52), local recurrence (p = 0.68), or late complications (rectal; p = 0.7, bladder; p = 0.95 or small intestine; p = 0.06) rates as compared to LDR brachytherapy. In the subgroup analysis no difference was observed for overall mortality and local recurrence in patients with clinical stages I, II and III. The quality of evidence was low for mortality and local recurrence in patients with clinical stage I, and moderate for other clinical stages. Conclusion Our meta-analysis shows that there are no differences between HDR and LDR for overall survival, local recurrence and late complications for clinical stages I, II and III. By means of the GRADE system, we recommend the use of HDR for all clinical stages of cervix cancer. PMID:19344527

  5. Gynecological brachytherapy - from low-dose-rate to high-tech. Gynaekologische Brachytherapie - von Low-dose-rate zu High-tech

    Energy Technology Data Exchange (ETDEWEB)

    Herrmann, T. (Abt. Strahlenthgerapie, Klinik und Poliklinik fuer Radiologie, Medizinische Akademie ' Carl Gustav Carus' , Dresden (Germany)); Christen, N. (Abt. Strahlenthgerapie, Klinik und Poliklinik fuer Radiologie, Medizinische Akademie ' Carl Gustav Carus' , Dresden (Germany)); Alheit, H.D. (Abt. Strahlenthgerapie, Klinik und Poliklinik fuer Radiologie, Medizinische Akademie ' Carl Gustav Carus' , Dresden (Germany))

    1993-03-01

    The transition from low-dose-rate (LDR) brachytherapy to high-dose-rate (HDR) afterloading treatment is in progress in most centres of radiation therapy. First reports of studies comparing HDR and LDR treatment in cervix cancer demonstrate nearly equal local control. In our own investigations on 319 patients with primary irradiated carcinoma of the cervix (125 HDR/194 LDR) we found the following control rates: Stage FIGO I 95.4%/82.9% (HDR versus LDR), stage FIGO II 71.4%/73.7%, stage FIGO III 57.9%/38.5%. The results are not significant. The side effects - scored after EORT/RTOG criteria - showed no significant differences between both therapies for serious radiogenic late effects on intestine, bladder and vagina. The study and findings from the literature confirm the advantage of the HDR-procedure for patient and radiooncologist and for radiation protection showing at least the same results as in the LDR-area. As for radiobiolgical point of view it is important to consider that the use of fractionation in the HDR-treatment is essential for the sparing of normal tissues and therefore a greater number of small fractionation doses in the brachytherapy should be desirable too. On the other hand the rules, which are true for fractionated percutaneous irradiation therapy (overall treatment time as short as possible to avoid reppopulation of tumor cells) should be taken into consideration in combined brachy-teletherapy regime in gynecologic tumors. The first step in this direction may be accelerated regime with a daily application of both treatment procedures. The central blocking of the brachytherapy region from the whole percutaneous treatment target volume should be critically reflected, especially in the case of advanced tumors. (orig.)

  6. Impact of using linear optimization models in dose planning for HDR brachytherapy

    International Nuclear Information System (INIS)

    Holm, Aasa; Larsson, Torbjoern; Carlsson Tedgren, Aasa

    2012-01-01

    Purpose: Dose plans generated with optimization models hitherto used in high-dose-rate (HDR) brachytherapy have shown a tendency to yield longer dwell times than manually optimized plans. Concern has been raised for the corresponding undesired hot spots, and various methods to mitigate these have been developed. The hypotheses upon this work is based are (a) that one cause for the long dwell times is the use of objective functions comprising simple linear penalties and (b) that alternative penalties, as these are piecewise linear, would lead to reduced length of individual dwell times. Methods: The characteristics of the linear penalties and the piecewise linear penalties are analyzed mathematically. Experimental comparisons between the two types of penalties are carried out retrospectively for a set of prostate cancer patients. Results: When the two types of penalties are compared, significant changes can be seen in the dwell times, while most dose-volume parameters do not differ significantly. On average, total dwell times were reduced by 4.2%, with a reduction of maximum dwell times by 25%, when the alternative penalties were used. Conclusions: The use of linear penalties in optimization models for HDR brachytherapy is one cause for the undesired long dwell times that arise in mathematically optimized plans. By introducing alternative penalties, a significant reduction in dwell times can be achieved for HDR brachytherapy dose plans. Although various measures for mitigating the long dwell times are already available, the observation that linear penalties contribute to their appearance is of fundamental interest.

  7. Tumour alpha/beta ratios and dose-rate selection in brachytherapy

    International Nuclear Information System (INIS)

    Duchesne, G.M.

    2003-01-01

    Traditionally brachytherapy employed low dose rate (LDR) techniques. Recent adoption of high dose rate (HDR) applications, addressing radiation protection concerns, has sparked debate over possible reductions in therapeutic ratio. The radiobiological characteristics of two contrasting examples, prostate cancer and cervical cancer, are examined. Both in-vitro and clinical observations of prostate cancer suggest a low α/β ratio. Labelling indices are below 2.5%, translating into long potential doubling times (Tpot ) of 16 to 61 days or more. Clinical PSA doubling times are in the order of years. Analysis of clinical endpoints in prostate cancer treated with either LDR or HDR techniques indicates that its α/β ratio may lie between 1 - 4 Gy, similar to slowly proliferating late reacting tissues. As such, therapeutic gain may arise from the use of hypofractionated HDR treatments, exploiting the sensitivity to large fraction sizes, effectively escalating dose. The slow proliferative rate also gives credence to the use of LDR, although several tumour doublings may occur during the effective treatment time, and analysis of the clinical data using a low α/β ratio suggests that LDR doses are only equivalent to 70 Gy with conventional fractionation. Cervical carcinoma is a rapidly proliferating tumour with Tpot values of 3-6 days. LDR implants were delivered over relatively short treatment times, negating repopulation effects, and the 'hyperfractionation' effect of LDR was suited to the high α/β ratio. HDR, although also preventing significant repopulation, has the potential to decrease the therapeutic ratio if low α/β , late-reacting tissues are not protected. Clinical data however show improved outcomes and reduced morbidity with HDR through reduced doses to normal tissues. Choosing the optimal dose rate in brachytherapy depends on tumour behaviour and achievable accuracy. HDR offers some advantages even for high α/β ratio tumours, and may be the technique of

  8. SU-E-T-169: Characterization of Pacemaker/ICD Dose in SAVI HDR Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Kalavagunta, C; Lasio, G; Yi, B; Zhou, J; Lin, M [Univ. of Maryland School Of Medicine, Baltimore, MD (United States)

    2015-06-15

    Purpose: It is important to estimate dose to pacemaker (PM)/Implantable Cardioverter Defibrillator (ICD) before undertaking Accelerated Partial Breast Treatment using High Dose Rate (HDR) brachytherapy. Kim et al. have reported HDR PM/ICD dose using a single-source balloon applicator. To the authors knowledge, there have so far not been any published PM/ICD dosimetry literature for the Strut Adjusted Volume Implant (SAVI, Cianna Medical, Aliso Viejo, CA). This study aims to fill this gap by generating a dose look up table (LUT) to predict maximum dose to the PM/ICD in SAVI HDR brachytherapy. Methods: CT scans for 3D dosimetric planning were acquired for four SAVI applicators (6−1-mini, 6−1, 8−1 and 10−1) expanded to their maximum diameter in air. The CT datasets were imported into the Elekta Oncentra TPS for planning and each applicator was digitized in a multiplanar reconstruction window. A dose of 340 cGy was prescribed to the surface of a 1 cm expansion of the SAVI applicator cavity. Cartesian coordinates of the digitized applicator were determined in the treatment leading to the generation of a dose distribution and corresponding distance-dose prediction look up table (LUT) for distances from 2 to 15 cm (6-mini) and 2 to 20 cm (10–1).The deviation between the LUT doses and the dose to the cardiac device in a clinical case was evaluated. Results: Distance-dose look up table were compared to clinical SAVI plan and the discrepancy between the max dose predicted by the LUT and the clinical plan was found to be in the range (−0.44%, 0.74%) of the prescription dose. Conclusion: The distance-dose look up tables for SAVI applicators can be used to estimate the maximum dose to the ICD/PM, with a potential usefulness for quick assessment of dose to the cardiac device prior to applicator placement.

  9. MO-C-17A-11: A Segmentation and Point Matching Enhanced Deformable Image Registration Method for Dose Accumulation Between HDR CT Images

    International Nuclear Information System (INIS)

    Zhen, X; Chen, H; Zhou, L; Yan, H; Jiang, S; Jia, X; Gu, X; Mell, L; Yashar, C; Cervino, L

    2014-01-01

    Purpose: To propose and validate a novel and accurate deformable image registration (DIR) scheme to facilitate dose accumulation among treatment fractions of high-dose-rate (HDR) gynecological brachytherapy. Method: We have developed a method to adapt DIR algorithms to gynecologic anatomies with HDR applicators by incorporating a segmentation step and a point-matching step into an existing DIR framework. In the segmentation step, random walks algorithm is used to accurately segment and remove the applicator region (AR) in the HDR CT image. A semi-automatic seed point generation approach is developed to obtain the incremented foreground and background point sets to feed the random walks algorithm. In the subsequent point-matching step, a feature-based thin-plate spline-robust point matching (TPS-RPM) algorithm is employed for AR surface point matching. With the resulting mapping, a DVF characteristic of the deformation between the two AR surfaces is generated by B-spline approximation, which serves as the initial DVF for the following Demons DIR between the two AR-free HDR CT images. Finally, the calculated DVF via Demons combined with the initial one serve as the final DVF to map doses between HDR fractions. Results: The segmentation and registration accuracy are quantitatively assessed by nine clinical HDR cases from three gynecological cancer patients. The quantitative results as well as the visual inspection of the DIR indicate that our proposed method can suppress the interference of the applicator with the DIR algorithm, and accurately register HDR CT images as well as deform and add interfractional HDR doses. Conclusions: We have developed a novel and robust DIR scheme that can perform registration between HDR gynecological CT images and yield accurate registration results. This new DIR scheme has potential for accurate interfractional HDR dose accumulation. This work is supported in part by the National Natural ScienceFoundation of China (no 30970866 and no

  10. Results of combined photodynamic therapy (PDT) and high dose rate brachytherapy (HDR) in treatment of obstructive endobronchial non-small cell lung cancer

    Science.gov (United States)

    Weinberg, Benjamin D.; Allison, Ron R.; Sibata, Claudio; Parent, Teresa; Downie, Gordon

    2009-06-01

    We reviewed the outcome of combined photodynamic therapy (PDT) and high dose rate brachytherapy (HDR) for patients with symptomatic obstruction from endobronchial non-small cell lung cancer. Methods: Nine patients who received combined PDT and HDR for endobronchial cancers were identified and their charts reviewed. The patients were eight males and one female aged 52-73 at diagnosis, initially presenting with various stages of disease: stage IA (N=1), stage IIA (N=1), stage III (N=6), and stage IV (N=1). Intervention was with HDR (500 cGy to 5 mm once weekly for 3 weeks) and PDT (2 mg/kg Photofrin, followed by 200 J/cm2 illumination 48 hours post infusion). Treatment group 1 (TG-1, N=7) received HDR first; Treatment group 2 (TG-2, N=2) received PDT first. Patients were followed by regular bronchoscopies. Results: Treatments were well tolerated, all patients completed therapy, and none were lost to follow-up. In TG-1, local tumor control was achieved in six of seven patients for: 3 months (until death), 15 months, 2+ years (until death), 2+ years (ongoing), and 5+ years (ongoing, N=2). In TG-2, local control was achieved in only one patient, for 84 days. Morbidities included: stenosis and/or other reversible benign local tissue reactions (N=8); photosensitivity reaction (N=2), and self-limited pleural effusion (N=2). Conclusions: Combined HDR/PDT treatment for endobronchial tumors is well tolerated and can achieve prolonged local control with acceptable morbidity when PDT follows HDR and when the spacing between treatments is one month or less. This treatment regimen should be studied in a larger patient population.

  11. SU-F-T-65: AutomaticTreatment Planning for High-Dose Rate (HDR) Brachytherapy with a VaginalCylinder Applicator

    International Nuclear Information System (INIS)

    Zhou, Y; Tan, J; Jiang, S; Albuquerque, K; Jia, X

    2016-01-01

    Purpose: High dose rate (HDR) brachytherapy treatment planning is conventionally performed in a manual fashion. Yet it is highly desirable to perform computerized automated planning to improve treatment planning efficiency, eliminate human errors, and reduce plan quality variation. The goal of this research is to develop an automatic treatment planning tool for HDR brachytherapy with a cylinder applicator for vaginal cancer. Methods: After inserting the cylinder applicator into the patient, a CT scan was acquired and was loaded to an in-house developed treatment planning software. The cylinder applicator was automatically segmented using image-processing techniques. CTV was generated based on user-specified treatment depth and length. Locations of relevant points (apex point, prescription point, and vaginal surface point), central applicator channel coordinates, and dwell positions were determined according to their geometric relations with the applicator. Dwell time was computed through an inverse optimization process. The planning information was written into DICOM-RT plan and structure files to transfer the automatically generated plan to a commercial treatment planning system for plan verification and delivery. Results: We have tested the system retrospectively in nine patients treated with vaginal cylinder applicator. These cases were selected with different treatment prescriptions, lengths, depths, and cylinder diameters to represent a large patient population. Our system was able to generate treatment plans for these cases with clinically acceptable quality. Computation time varied from 3–6 min. Conclusion: We have developed a system to perform automated treatment planning for HDR brachytherapy with a cylinder applicator. Such a novel system has greatly improved treatment planning efficiency and reduced plan quality variation. It also served as a testbed to demonstrate the feasibility of automatic HDR treatment planning for more complicated cases.

  12. SU-F-T-65: AutomaticTreatment Planning for High-Dose Rate (HDR) Brachytherapy with a VaginalCylinder Applicator

    Energy Technology Data Exchange (ETDEWEB)

    Zhou, Y; Tan, J; Jiang, S; Albuquerque, K; Jia, X [UT Southwestern Medical Center, Dallas, TX (United States)

    2016-06-15

    Purpose: High dose rate (HDR) brachytherapy treatment planning is conventionally performed in a manual fashion. Yet it is highly desirable to perform computerized automated planning to improve treatment planning efficiency, eliminate human errors, and reduce plan quality variation. The goal of this research is to develop an automatic treatment planning tool for HDR brachytherapy with a cylinder applicator for vaginal cancer. Methods: After inserting the cylinder applicator into the patient, a CT scan was acquired and was loaded to an in-house developed treatment planning software. The cylinder applicator was automatically segmented using image-processing techniques. CTV was generated based on user-specified treatment depth and length. Locations of relevant points (apex point, prescription point, and vaginal surface point), central applicator channel coordinates, and dwell positions were determined according to their geometric relations with the applicator. Dwell time was computed through an inverse optimization process. The planning information was written into DICOM-RT plan and structure files to transfer the automatically generated plan to a commercial treatment planning system for plan verification and delivery. Results: We have tested the system retrospectively in nine patients treated with vaginal cylinder applicator. These cases were selected with different treatment prescriptions, lengths, depths, and cylinder diameters to represent a large patient population. Our system was able to generate treatment plans for these cases with clinically acceptable quality. Computation time varied from 3–6 min. Conclusion: We have developed a system to perform automated treatment planning for HDR brachytherapy with a cylinder applicator. Such a novel system has greatly improved treatment planning efficiency and reduced plan quality variation. It also served as a testbed to demonstrate the feasibility of automatic HDR treatment planning for more complicated cases.

  13. Overview of Radiosensitivity of Human Tumor Cells to Low-Dose-Rate Irradiation

    International Nuclear Information System (INIS)

    Williams, Jerry R.; Zhang Yonggang; Zhou Haoming; Gridley, Daila S.; Koch, Cameron J.; Slater, James M.; Little, John B.

    2008-01-01

    Purpose: We compared clonogenic survival in 27 human tumor cell lines that vary in genotype after low-dose-rate (LDR) or high-dose rate (HDR) irradiation. We measured susceptibility to LDR-induced redistribution in the cell cycle in eight of these cell lines. Methods and Materials: We measured clonogenic survival after up to 96 hours of LDR (0.25 Gy/h) irradiation. We compared these with clonogenic survival after HDR irradiation (50 Gy/h). Using flow cytometry, we measured LDR-induced redistribution as a function of time during LDR irradiation in eight of these cell lines. Results: Coefficients that describe clonogenic survival after both LDR and HDR irradiation segregate into four radiosensitivity groups that associate with cell genotype: mutant (mut)ATM, wild-type TP53, mutTP53, and an unidentified gene in radioresistant glioma cells. The LDR and HDR radiosensitivity correlates at lower doses (∼2 Gy HDR, ∼6 Gy LDR), but not at higher doses (HDR > 4 Gy; LDR > 6 Gy). The rate of LDR-induced loss of clonogenic survival changes at approximately 24 hours; wild-type TP53 cells become more resistant and mutTP53 cells become more sensitive. Redistribution induced by LDR irradiation also changes at approximately 24 hours. Conclusions: Radiosensitivity of human tumor cells to both LDR and HDR irradiation is genotype dependent. Analysis of coefficients that describe cellular radiosensitivity segregates 27 cell lines into four statistically distinct groups, each associating with specific genotypes. Changes in cellular radiosensitivity and redistribution in the cell cycle are strongly time dependent. Our data establish a genotype-dependent time-dependent model that predicts clonogenic survival, explains the inverse dose-rate effect, and suggests possible clinical applications

  14. Gamma Low-Dose-Rate Ionizing Radiation Stimulates Adaptive Functional and Molecular Response in Human Aortic Endothelial Cells in a Threshold-, Dose-, and Dose Rate-Dependent Manner.

    Science.gov (United States)

    Vieira Dias, Juliana; Gloaguen, Celine; Kereselidze, Dimitri; Manens, Line; Tack, Karine; Ebrahimian, Teni G

    2018-01-01

    A central question in radiation protection research is whether low-dose and low-dose-rate (LDR) exposures to ionizing radiation play a role in progression of cardiovascular disease. The response of endothelial cells to different LDR exposures may help estimate risk of cardiovascular disease by providing the biological mechanism involved. We investigated the effect of chronic LDR radiation on functional and molecular responses of human aorta endothelial cells (HAoECs). Human aorta endothelial cells were continuously irradiated at LDR (6 mGy/h) for 15 days and analyzed at time points when the cumulative dose reached 0.05, 0.5, 1.0, and 2.0 Gy. The same doses were administered acutely at high-dose rate (HDR; 1 Gy/min). The threshold for the loss of angiogenic capacity for both LDR and HDR radiations was between 0.5 and 1.0 Gy. At 2.0 Gy, angiogenic capacity returned to normal only for HAoEC exposed to LDR radiation, associated with increased expression of antioxidant and anti-inflammatory genes. Pre-LDR, but not pre-HDR, radiation, followed by a single acute 2.0 Gy challenge dose sustained the expression of antioxidant and anti-inflammatory genes and stimulated angiogenesis. Our results suggest that dose rate is important in cellular response and that a radioadaptive response is involved for a 2.0 Gy dose at LDR.

  15. High-dose-rate brachytherapy alone post-hysterectomy for endometrial cancer

    International Nuclear Information System (INIS)

    MacLeod, Craig; Fowler, Allan; Duval, Peter; D'Costa, Ieta; Dalrymple, Chris; Firth, Ian; Elliott, Peter; Atkinson, Ken; Carter, Jonathan

    1998-01-01

    Purpose: To evaluate the outcome of post-hysterectomy adjuvant vaginal high-dose-rate (HDR) brachytherapy. Methods and Materials: A retrospective analysis was performed on a series of 143 patients with endometrial cancer treated with HDR brachytherapy alone post-hysterectomy from 1985 to June 1993. Of these patients, 141 received 34 Gy in four fractions prescribed to the vaginal mucosa in a 2-week period. The median follow-up was 6.9 years. Patients were analyzed for treatment parameters, survival, local recurrence, distant relapse, and toxicity. Results: Five-year relapse free survival and overall survival was 100% and 88% for Stage 1A, 98% and 94% for Stage IB, 100% and 86% for Stage IC, and 92% and 92% for Stage IIA. The overall vaginal recurrence rate was 1.4%. The overall late-toxicity rate was low, and no RTOG grade 3, 4, or 5 complications were recorded. Conclusion: These results are similar to reported international series that have used either low-dose-rate or HDR brachytherapy. The biological effective dose was low for both acute and late responding tissues compared with some of the HDR brachytherapy series, and supports using this lower dose and possibly decreasing late side-effects with no apparent increased risk of vaginal recurrence

  16. Toward endobronchial Ir-192 high-dose-rate brachytherapy therapeutic optimization

    International Nuclear Information System (INIS)

    Gay, H A; Allison, R R; Downie, G H; Mota, H C; Austerlitz, C; Jenkins, T; Sibata, C H

    2007-01-01

    A number of patients with lung cancer receive either palliative or curative high-dose-rate (HDR) endobronchial brachytherapy. Up to a third of patients treated with endobronchial HDR die from hemoptysis. Rather than accept hemoptysis as an expected potential consequence of HDR, we have calculated the radial dose distribution for an Ir-192 HDR source, rigorously examined the dose and prescription points recommended by the American Brachytherapy Society (ABS), and performed a radiobiological-based analysis. The radial dose rate of a commercially available Ir-192 source was calculated with a Monte Carlo simulation. Based on the linear quadratic model, the estimated palliative, curative and blood vessel rupture radii from the center of an Ir-192 source were obtained for the ABS recommendations and a series of customized HDR prescriptions. The estimated radius at risk for blood vessel perforation for the ABS recommendations ranges from 7 to 9 mm. An optimized prescription may in some situations reduce this radius to 4 mm. The estimated blood perforation radius is generally smaller than the palliative radius. Optimized and individualized endobronchial HDR prescriptions are currently feasible based on our current understanding of tumor and normal tissue radiobiology. Individualized prescriptions could minimize complications such as fatal hemoptysis without sacrificing efficacy. Fiducial stents, HDR catheter centering or spacers and the use of CT imaging to better assess the relationship between the catheter and blood vessels promise to be useful strategies for increasing the therapeutic index of this treatment modality. Prospective trials employing treatment optimization algorithms are needed

  17. Late effects of post-high-dose-rate brachytherapy for oropharyngeal carcinoma: are they severer than post-low-dose-rate?

    International Nuclear Information System (INIS)

    Nose, T.; Koizumi, M.; Nishiyama, K.; Peiffert, D.; Lapeyre, M.; Hoffstetter, S.

    2004-01-01

    Background: late effects by high-dose-rate (HDR) brachytherapy have been believed severer than low-dose-rate (LDR) provided tumor control was constant. Local control of oropharyngeal carcinoma with HDR at Osaka Medical Center was comparable to LDR series from Centre Alexis Vautrin (82%, 79.5%, respectively). To assess the feasibility of HDR brachytherapy, the late effects were compared. Patients and methods: the data of 29 HDR and 24 LDR patients (median follow-up of 27 and 29.5 months, respectively; p = 0.89) were collected. The HDR schedule was 21 Gy/3.5 fractions/2 days following 46 Gy/23 fractions external beam, while 25 Gy/3 days following 50 Gy/25 fractions external beam was for LDR. Late changes were evaluated using RTOG/EORTC late morbidity scoring scheme. For subclinical late changes, mucosa chapter of Dische score was modified for brachytherapy. Scores were discussed through photos and were agreed on by authors. Late sequelae were estimated, by reviewing charts, concerning frequency, severity, and duration of mucosal damages (erosion and ulcer). Results: Late changes were of no difference (p = 0.12 for EORTC/RTOG, and p = 0.45, 0.47, 1.00, 0.12, 0.16, 0.95, 0.27, 0.21 for erythema, ulceration, edema, thinning, pallor, telangiectasia, mobility impairment of tongue/faucial pillars, respectively, of the modified Dische score). Late sequelae showed no differences (p = 0.90, 0.12, 0.40 for frequency, severity, duration, respectively, of mucosal damages). Conclusion: the late effects by HDR were not severer than by LDR. HDR oropharyngeal brachytherapy is as safe as LDR. (orig.)

  18. Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source

    International Nuclear Information System (INIS)

    Sarabiasl, Akbar; Ayoobian, Navid; Jabbari, Iraj; Poorbaygi, Hossein; Javanshir, Mohammad Reza

    2016-01-01

    High-dose-rate (HDR) brachytherapy is a common method for cancer treatment in clinical brachytherapy. Because of the different source designs, there is a need for specific dosimetry data set for each HDR model. The purpose of this study is to obtain detailed dose rate distributions in water phantom for a first prototype HDR 192 Ir brachytherapy source model, IRAsource, and compare with the other published works. In this study, Monte Carlo N-particle (MCNP version 4C) code was used to simulate the dose rate distributions around the HDR source. A full set of dosimetry parameters reported by the American Association of Physicists in Medicine Task Group No. 43U1 was evaluated. Also, the absorbed dose rate distributions in water, were obtained in an along-away look-up table. The dose rate constant, Λ, of the IRAsource was evaluated to be equal to 1.112 ± 0.005 cGy h −1 U −1 . The results of dosimetry parameters are presented in tabulated and graphical formats and compared with those reported from other commercially available HDR 192 Ir sources, which are in good agreement. This justifies the use of specific data sets for this new source. The results obtained in this study can be used as input data in the conventional treatment planning systems.

  19. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic

    International Nuclear Information System (INIS)

    Palmer, A. L.; Di Pietro, P.; Alobaidli, S.; Issa, F.; Doran, S.; Bradley, D.; Nisbet, A.

    2013-01-01

    Purpose: Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE ® with optical-CT readout. Methods: Ge-doped SiO 2 fibers with 6 μm active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 μm active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE ® , 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. Results: All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE ® , and the latter two suffered from unacceptable noise and artifact. For the experimental conditions used in this study, the useful range from an isolated

  20. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic.

    Science.gov (United States)

    Palmer, A L; Di Pietro, P; Alobaidli, S; Issa, F; Doran, S; Bradley, D; Nisbet, A

    2013-06-01

    Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE(®) with optical-CT readout. Ge-doped SiO2 fibers with 6 μm active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 μm active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE(®), 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE(®), and the latter two suffered from unacceptable noise and artifact. For the experimental conditions used in this study, the useful range from an isolated HDR source was 5-40 mm for

  1. High versus low dose-rate intracavitary irradiation for adenocarcinoma of the uterine cervix

    International Nuclear Information System (INIS)

    Kim, Woo Chul; Kim, Gwi Eon; Chung, Eun Ji; Suh, Chang Ok; Hong, Soon Won; Cho, Young Kap; Loh, John Jk

    1999-01-01

    The incidence of adenocarcinoma of the uterine cervix is low. Traditionally, Low Dose Rate (LDR) brachytherapy has been used as a standard modality in the treatment for patients with carcinoma of the uterine cervix. The purpose of this report is to evaluate the effects of the High Dose Rate (HDR) brachytherapy in the patients with adenocarcinoma of the uterine cervix compared with the LDR. From January 1971 to December 1992, 106 patients of adenocarcinoma of uterine cervix were treated with radiation therapy in the Department of Radiation Oncology, Yonsei University with curative intent. LDR brachytherapy was carried out on 35 patients and 71 patients were treated with HDR brachytherapy. In LDR Group, 8 patients were in stage I, 18 in stage II and 9 in stage III. external radiation therapy was delivered with 10 MV X-ray, daily 2 Gy fractionation, total dose 40-46 Gy (median 48 Gy). And LDR Radium intracavitary irradiation was performed with Henschke applicator, 22-56 Gy to point A (median 43 Gy). In HDR Group, there were 16 patients in stage I, 38 in stage II and 17 in stage III. The total dose of external radiation was 40-61 Gy (median 45 Gy), daily 1.8-2.0 Gy. HDR Co-60 intracavitary irradiation was performed with RALS(Remote Afterloading System), 30-57 Gy (median 39 Gy) to point A, 3 times a week, 3 Gy per fraction. The 5-year overall survival rate in LDR Group was 72.9%, 61.9%, 45.0% in stage I, II, III, respectively and corresponding figures for HDR were 87.1%, 58.3%, 41.2%, respectively (p>0.05). There was no statistical difference in terms of the 5-year overall survival rate between HDR Group and LDR Group in adenocarcinoma of the uterine cervix. There was 11% of late complication rates in LDR Group and 27% in HDR Group. There were no prognostic factors compared HDR with LDR group. The incidence of the late complication rate in HDR Group stage II, III was higher than that in LDR Group (16.7% vs. 31.6% in stage II, 11.1% vs. 35.3% in stage III, p>0

  2. Response of mouse lung to irradiation at different dose-rates

    International Nuclear Information System (INIS)

    Hill, R.P.

    1983-01-01

    Groups of LAF1 mice were given thoracic irradiation using 60 Co γ-rays at dose-rates of 0.05 Gy/min (LDR) or 1.1 Gy/min (HDR) and the death of the animals was monitored as a function of time. It was found that the time pattern of animal deaths was similar for the two different dose-rates. Dose response curves for animals dying at various times up to 500 days after irradiation were calculated and the LD 50 values determined. The curves for the LD 50 values, plotted as a function of the time at analysis for treatment at HDR or LDR, were essentially parallel to each other but separated by a factor (LDR/HDR) of about 1.8. This indicates that the sparing effect of LDR treatment is the same for deaths occurring during the early pneumonitis phase or during the late fibrotic phase of lung damage. The available information on the response of patients to whole thoracic irradiation, given for either palliation or piror to bone marrow transplantation, suggests that for similar dose-rates to those studied here the ratio (LDR/HDR) is only 1.2 to 1.3. This difference between the animal and human data may reflect the modifying effect of the large doses of cytotoxic drugs used in combination with the irradiation of bone marrow transplant patients

  3. Late change of normal tissue treated either by high dose rate or low dose rate interstitial brachytherapy. A retrospective comparative study on oral and oropharyngeal mucosa

    International Nuclear Information System (INIS)

    Nose, Takayuki; Koizumi, Masahiko; Nishiyama, Kinji; Inoue, Toshihiko

    2002-01-01

    The purpose of this study was to compare late changes of normal tissue treated either by high dose rate (HDR) or low dose rate (LDR) interstitial brachytherapy. For HDR group, 22 oropharynx cancer patients who were treated by HDR Ir-192 interstitial brachytherapy with/without external beam radiotherapy in Osaka (Osaka Medical Center for Cancer and Cardiovascular Diseases and Osaka University Hospital) during June 1994 through April 2000 and came to the follow-up clinics during July 2000 through December 2000 were studied. For LDR group, 26 oropharynx cancer patients who were treated by LDR Ir-192 interstitial brachytherapy with/without external beam radiotherapy in Nancy (Centre Alexis Vautrin) during February 1989 through July 1998 and came to the follow-up clinics during April 1999 through July 1999 were studied. The standard HDR schedules were 54 Gy/9 fr/5-6 days for monotherapy and 18-24 Gy/3-4 fr/2-3 days following 45 Gy external beam radiotherapy. The standard LDR schedules were 65 Gy/5-6 days for monotherapy and 15-25 Gy/2-3 days following 50 Gy external beam radiotherapy. For evaluation of the late changes, we scored the mucosal and muscular changes inside the treated volume using the modified Dische score system and the RTOG/EORTC late radiation morbidity scoring scheme. For 6 items of the modified Dische score system, no significant difference was found between HDR and LDR groups. For the remaining 2 items (pallor, mobility impairment of faucial pillars), LDR group showed higher scores (p=0.010, 0.002). LDR group showed a trend toward higher scores for the RTOG/EORTC scheme (p=0.059). Some predict late effects by HDR interstitial brachytherapy to be severer than by LDR because no dose-rate effects can be expected. Our study, however, showed at least equivalent or even milder late changes by HDR. Appropriate fractionation schedule and extra geometrical sparing effects by optimized dose distribution of HDR group might result in milder late changes. With our

  4. High-dose-rate interstitial brachytherapy for the treatment of penile carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Petera, J.; Odrazka, K.; Zouhar, M.; Bedrosova, J.; Dolezel, M. [Dept. of Oncology and Radiotherapy, Charles Univ. Medical School and Teaching Hospital, Hradec Kralove (Czech Republic)

    2004-02-01

    Background: interstitial low-dose-rate (LDR) brachytherapy allows conservative treatment of T1-T2 penile carcinoma. High-dose-rate (HDR) is often considered to be dangerous for interstitial implants because of a higher risk of complications, but numerous reports suggest that results may be comparable to LDR. Nevertheless, there are no data in the literature available regarding HDR interstitial brachytherapy for carcinoma of the penis. Case report: a 64-year-old man with T1 NO MO epidermoid carcinoma of the glans is reported. Interstitial HDR brachytherapy was performed using the stainless hollow needle technique and a breast template for fixation and good geometry. The dose delivered was 18 x 3 Gy twice daily. Results: after 232 days from brachytherapy, the patient was without any evidence of the tumor, experienced no serious radiation-induced complications, and had a fully functional organ. Conclusion: HDR interstitial brachytherapy is feasible in selected case of penis carcinoma, when careful planning and small single fractions are used. (orig.)

  5. Development of optimized dosimetric models for HDR brachytherapy

    International Nuclear Information System (INIS)

    Thayalan, K.; Jagadeesan, M.

    2003-01-01

    High dose rate brachytherapy (HDRB) systems are in clinical use for more than four decades particularly in cervical cancer. Optimization is the method to produce dose distribution which assures that doses are not compromised at the treatment sites whilst reducing the risk of overdosing critical organs. Hence HDRB optimization begins with the desired dose distribution and requires the calculations of the relative weighting factors for each dwell position with out changing the source activity. The optimization for Ca. uterine cervix treatment is simply duplication of the dose distribution used for Low dose rate (LDR) applications. In the present work, two optimized dosimetric models were proposed and studied thoroughly, to suit the local clinical conditions. These models are named as HDR-C and HDR-D, where C and D represent configuration and distance respectively. These models duplicate exactly the LDR pear shaped dose distribution, which is a golden standard. The validity of these models is tested in different clinical situations and in actual patients (n=92). These models: HDR-C and HDR-D reduce bladder dose by 11.11% and 10% and rectal dose by 8% and 7% respectively. The treatment time is also reduced by 12-14%. In a busy hospital setup, these models find a place to cater large number of patients, while addressing individual patients geometry. (author)

  6. Dose escalation using conformal high-dose-rate brachytherapy improves outcome in unfavorable prostate cancer.

    Science.gov (United States)

    Martinez, Alvaro A; Gustafson, Gary; Gonzalez, José; Armour, Elwood; Mitchell, Chris; Edmundson, Gregory; Spencer, William; Stromberg, Jannifer; Huang, Raywin; Vicini, Frank

    2002-06-01

    To overcome radioresistance for patients with unfavorable prostate cancer, a prospective trial of pelvic external beam irradiation (EBRT) interdigitated with dose-escalating conformal high-dose-rate (HDR) prostate brachytherapy was performed. Between November 1991 and August 2000, 207 patients were treated with 46 Gy pelvic EBRT and increasing HDR brachytherapy boost doses (5.50-11.5 Gy/fraction) during 5 weeks. The eligibility criteria were pretreatment prostate-specific antigen level >or=10.0 ng/mL, Gleason score >or=7, or clinical Stage T2b or higher. Patients were divided into 2 dose levels, low-dose biologically effective dose 93 Gy (149 patients). No patient received hormones. We used the American Society for Therapeutic Radiology and Oncology definition for biochemical failure. The median age was 69 years. The mean follow-up for the group was 4.4 years, and for the low and high-dose levels, it was 7.0 and 3.4 years, respectively. The actuarial 5-year biochemical control rate was 74%, and the overall, cause-specific, and disease-free survival rate was 92%, 98%, and 68%, respectively. The 5-year biochemical control rate for the low-dose group was 52%; the rate for the high-dose group was 87% (p failure. The Radiation Therapy Oncology Group Grade 3 gastrointestinal/genitourinary complications ranged from 0.5% to 9%. The actuarial 5-year impotency rate was 51%. Pelvic EBRT interdigitated with transrectal ultrasound-guided real-time conformal HDR prostate brachytherapy boost is both a precise dose delivery system and a very effective treatment for unfavorable prostate cancer. We demonstrated an incremental beneficial effect on biochemical control and cause-specific survival with higher doses. These results, coupled with the low risk of complications, the advantage of not being radioactive after implantation, and the real-time interactive planning, define a new standard for treatment.

  7. Guidelines for optimization of planar HDR implants

    International Nuclear Information System (INIS)

    Zwicker, R.D.; Schmidt-Ullrich, R.

    1996-01-01

    Purpose: Conventional low dose rate (LDR) planar Ir-192 implants are typically carried out using at most a few different source strengths. Remote afterloading offers a much higher degree of flexibility with individually programmable dwell times. Dedicated software is available to generate individual dwell times producing isodose surfaces which contour as closely as possible the target volume. The success of these algorithms in enclosing the target volume while sparing normal tissues is dependent on the positioning of the source guides which constrain the dwell points. In this work we provide source placement guidelines for optimal coverage and dose uniformity in planar high dose rate (HDR) implants. The resulting distributions are compared with LDR treatments in terms of dose uniformity and early and late tissue effects. Materials and methods: Computer studies were undertaken to determine source positions and dwell times for optimal dose uniformity in planar HDR implants, and the results were compared to those obtained using corresponding LDR implant geometries. The improvements in the dose distributions achieved with the remote after loader are expected to help offset the increased late tissue effects which can occur when LDR irradiation is replaced with a few large HDR fractions. Equivalent differential volume-dose (DVD) curves for early and late effects were calculated for different numbers of HDR fractions using a linear-quadratic model and compared to the corresponding curves for the LDR regime. Results: Tables of source placement parameters were generated as guidelines for achieving highly homogeneous planar HDR dose distributions. Differential volume-dose data generated inside the target volume provide a quantitative measure of the improvement in real dose homogeneity obtained with remote afterloading. The net result is a shift of the peak in the DVD curve toward lower doses relative to the LDR implant. The equivalent DVD curves for late effects obtained

  8. The status of low dose rate and future of high dose rate Cf-252 brachytherapy

    International Nuclear Information System (INIS)

    Rivard, M.J.; Wierzbicki, J.G.; Van den Heuvel, F.; Chuba, P.J.; Fontanesi, J.

    1997-12-01

    This work describes the current status of the US low dose rate (LDR) Cf-252 brachytherapy program. The efforts undertaken towards development of a high dose rate (HDR) remotely after loaded Cf-252 source, which can accommodate 1 mg or greater Cf-252, are also described. This HDR effort is a collaboration between Oak Ridge National Laboratory (ORNL), commercial remote after loader manufactures, the Gershenson Radiation Oncology Center (ROC), and Wayne State University. To achieve this goal, several advances in isotope chemistry and source preparation at ORNL must be achieved to yield a specific material source loading of greater than or equal 1 mg Cf-252 per mm3. Development work with both radioactive and non-radioactive stand-ins for Cf-252 have indicated the feasibility of fabricating such sources. As a result, the decreased catheter diameter and computer controlled source placement will permit additional sites (e.g. brain, breast, prostate, lung, parotid, etc.) to be treated effectively with Cf-252 sources. Additional work at the Radiochemical Engineering and Development Center (REDC) remains in source fabrication, after loader modification, and safe design. The current LDR Cf-252 Treatment Suite at the ROC is shielded and licensed to hold up to 1 mg of Cf-252. This was designed to maintain cumulative personnel exposure, both external to the room and in direct isotope handling, at less than 20 microSv/hr. However, cumulative exposure may be greatly decreased if a Cf-252 HDR unit is employed which would eliminate direct isotope handling and decrease treatment times from tilde 3 hours to an expected range of 3 to 15 minutes. Such a Cf-252 HDR source will also demonstrate improved dose distributions over current LDR treatments due to the ability to step the point-like source throughout the target volume and weight the dwell time accordingly

  9. Rectal Bleeding After High-Dose-Rate Brachytherapy Combined With Hypofractionated External-Beam Radiotherapy for Localized Prostate Cancer: The Relationship Between Dose-Volume Histogram Parameters and the Occurrence Rate

    International Nuclear Information System (INIS)

    Okamoto, Masahiko; Ishikawa, Hitoshi; Ebara, Takeshi; Kato, Hiroyuki; Tamaki, Tomoaki; Akimoto, Tetsuo; Ito, Kazuto; Miyakubo, Mai; Yamamoto, Takumi; Suzuki, Kazuhiro; Takahashi, Takeo; Nakano, Takashi

    2012-01-01

    Purpose: To determine the predictive risk factors for Grade 2 or worse rectal bleeding after high-dose-rate brachytherapy (HDR-BT) combined with hypofractionated external-beam radiotherapy (EBRT) for prostate cancer using dose–volume histogram analysis. Methods and Materials: The records of 216 patients treated with HDR-BT combined with EBRT were analyzed. The treatment protocols for HDR-BT were 5 Gy × five times in 3 days or 7 Gy × three, 10.5 Gy × two, or 9 Gy × two in 2 days. The EBRT doses ranged from 45 to 51 Gy with a fractional dose of 3 Gy. Results: In 20 patients Grade 2 or worse rectal bleeding developed, and the cumulative incidence rate was 9% at 5 years. By converting the HDR-BT and EBRT radiation doses into biologic effective doses (BED), the BED 3 at rectal volumes of 5% and 10% in the patients who experienced bleeding were significantly higher than those in the remaining 196 patients. Univariate analysis showed that a higher rectal BED 3–5% and the use of fewer needles in brachytherapy were correlated with the incidence of bleeding, but BED 3–5% was found to be the only significant factor on multivariate analysis. Conclusions: The radiation dose delivered to small rectal lesions as 5% is important for predicting Grade 2 or worse rectal bleeding after HDR-BT combined with EBRT for prostate cancer.

  10. Endorectal high dose rate brachytherapy quality assurance

    International Nuclear Information System (INIS)

    Devic, S.; Vuong, T.; Evans, M.; Podgorsak, E.

    2008-01-01

    We describe our quality assurance method for preoperative high dose rate (HDR) brachytherapy of endorectal tumours. Reproduction of the treatment planning dose distribution on a daily basis is crucial for treatment success. Due to the cylindrical symmetry, two types of adjustments are necessary: applicator rotation and dose distribution shift along the applicator axis. (author)

  11. Prostate Specific Antigen (PSA) as Predicting Marker for Clinical Outcome and Evaluation of Early Toxicity Rate after High-Dose Rate Brachytherapy (HDR-BT) in Combination with Additional External Beam Radiation Therapy (EBRT) for High Risk Prostate Cancer.

    Science.gov (United States)

    Ecke, Thorsten H; Huang-Tiel, Hui-Juan; Golka, Klaus; Selinski, Silvia; Geis, Berit Christine; Koswig, Stephan; Bathe, Katrin; Hallmann, Steffen; Gerullis, Holger

    2016-11-10

    High-dose-rate brachytherapy (HDR-BT) with external beam radiation therapy (EBRT) is a common treatment option for locally advanced prostate cancer (PCa). Seventy-nine male patients (median age 71 years, range 50 to 79) with high-risk PCa underwent HDR-BT following EBRT between December 2009 and January 2016 with a median follow-up of 21 months. HDR-BT was administered in two treatment sessions (one week interval) with 9 Gy per fraction using a planning system and the Ir192 treatment unit GammaMed Plus iX. EBRT was performed with CT-based 3D-conformal treatment planning with a total dose administration of 50.4 Gy with 1.8 Gy per fraction and five fractions per week. Follow-up for all patients was organized one, three, and five years after radiation therapy to evaluate early and late toxicity side effects, metastases, local recurrence, and prostate-specific antigen (PSA) value measured in ng/mL. The evaluated data included age, PSA at time of diagnosis, PSA density, BMI (body mass index), Gleason score, D'Amico risk classification for PCa, digital rectal examination (DRE), PSA value after one/three/five year(s) follow-up (FU), time of follow-up, TNM classification, prostate volume, and early toxicity rates. Early toxicity rates were 8.86% for gastrointestinal, and 6.33% for genitourinary side effects. Of all treated patients, 84.81% had no side effects. All reported complications in early toxicity were grade 1. PSA density at time of diagnosis ( p = 0.009), PSA on date of first HDR-BT ( p = 0.033), and PSA on date of first follow-up after one year ( p = 0.025) have statistical significance on a higher risk to get a local recurrence during follow-up. HDR-BT in combination with additional EBRT in the presented design for high-risk PCa results in high biochemical control rates with minimal side-effects. PSA is a negative predictive biomarker for local recurrence during follow-up. A longer follow-up is needed to assess long-term outcome and toxicities.

  12. MO-B-BRC-02: Ultrasound Based Prostate HDR

    Energy Technology Data Exchange (ETDEWEB)

    Chang, Z. [Duke University Medical Center (United States)

    2016-06-15

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions.

  13. MO-B-BRC-02: Ultrasound Based Prostate HDR

    International Nuclear Information System (INIS)

    Chang, Z.

    2016-01-01

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions

  14. MO-B-BRC-04: MRI-Based Prostate HDR

    International Nuclear Information System (INIS)

    Mourtada, F.

    2016-01-01

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions

  15. MO-B-BRC-03: CT-Based Prostate HDR

    International Nuclear Information System (INIS)

    Zoberi, J.

    2016-01-01

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions

  16. MO-B-BRC-04: MRI-Based Prostate HDR

    Energy Technology Data Exchange (ETDEWEB)

    Mourtada, F. [Christiana Care Hospital (United States)

    2016-06-15

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions.

  17. MO-B-BRC-03: CT-Based Prostate HDR

    Energy Technology Data Exchange (ETDEWEB)

    Zoberi, J. [Washington University School of Medicine (United States)

    2016-06-15

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions.

  18. High dose rate versus low dose rate brachytherapy in the treatment of stage IIIB cervical cancer, and the importance of brachytherapy timing

    International Nuclear Information System (INIS)

    Petereit, Daniel G.; Sarkaria, Jann N.; Czyzewski, Ann; Buchler, Dolores A.

    1996-01-01

    Purpose/Objective: To determine the efficacy of HDR versus LDR brachytherapy for Stage IIIB cervical cancer patients. Material and Methods: Forty-four HDR patients were retrospectively compared to 51 LDR patients treated at the same institution from 1977 to 1988 (LDR) and from 1989 to 1995 (HDR). A tumor burden score (TBS) of 2-9 was calculated for both groups of patients to assess volume of disease (2-4 low tumor burden, 5-9 high tumor burden). LDR and HDR patients received 60 Gy to the whole pelvis at 1.7 Gy/Fx. The majority of LDR patients were treated after completion of external beam radiation (EBR) with one 25 Gy implant to Point A (55 cGy/h). The HDR patients were treated with 4-5 HDR fractions of 3.7 Gy to 5.8 Gy/Fx for an LDR equivalence of 20-25 Gy (median dose/Fx 4.3 Gy, median insertion number 5). Clinical endpoints were calculated actuarially with significance determined by the log rank test and the relative risk (RR). Results: The median follow-up for the HDR and LDR groups was 1.8 and 5 years, respectively. Pelvic control and survival was better in the LDR group than the HDR group, 51%, 73%, 32%, 44% (p = 0.004, RR = 0.4), with grade III and above RTOG complications of 19% and 15%, respectively. The median age and performance status were similar between the two groups; however, a TBS score ≥7 was present in 23% of the HDR patients and in 9% of the LDR patients. Pelvic control in the HDR group was 58% with a TBS ≤4, and 17% with a TBS >4 (p = 0.01, RR = 0.4). The median EBR dose at the first HDR insertion was 31 Gy while all the LDR patients received 60 Gy before the insertion. Pelvic control rates in Table 1 indicate a trend towards improved outcome within the HDR group and same TBS if more external beam radiation was given prior to the first HDR insertion. Pelvic control was also higher within the HDR group when Point A received a BED Gy 10≥100 versus <100: 62%, 40%, respectively (RR 0.6). Conclusion: These retrospective results of HDR versus

  19. HDR- and LDR-interstitial irradiation (IRT) in rat spinal cord: the effect of decreasing the dose rate and the impact of a rapid dose fall off over the spinal cord

    International Nuclear Information System (INIS)

    Pop, L.A.M.; Plas, M. van der; Hanssen, A.E.J.; Kogel, A.J. van der

    1996-01-01

    Introduction: Detailed knowledge of radiobiological parameters of the different tissues involved are warranted before HDR- and recently PDR-brachytherapy can be successfully introduced in clinical practice as an alternative to LDR- brachytherapy. The purpose of this study is to determine the α/β ratio and half time of repair of rat spinal cord during continuous irradiation at different dose rates and to investigate the impact of a rapid dose fall off over the spinal cord thickness. Material and methods: Two parallel catheters are inserted on each side of the vertebral bodies from the level of Th 10 to L 4 . These catheters were afterloaded with two 192 Ir- wires of 4 cm length each (activity 1- 10 mCi/cm) or connected to the HDR-microSelectron. Serial experiments have been carried out to obtain complete dose response curves at 5 different dose rates, resp. 0.5, 0.9, 1.6, 2.6 and 120 Gy/h. Paralysis of the hindlegs after 5-6 months and histopathological examination of the spinal cord of each animal are used as experimental endpoints. Dose-volume histograms of each irradiated rat have been analysed to evaluate the correlation between dose distribution and biological response and the histopathological damage seen. Results: The distribution of the histological damage was a good reflection of the rapid dose fall-off over the spinal cord, with white matter necrosis or demyelination predominantly seen in the dorsal tracts of the spinal cord or dorsal roots. With each reduction of the dose rate, spinal cord tolerance was significantly increased, with a maximum dose rate factor of 4.3 if the dose rate was reduced from 120 Gy/h to 0.53 Gy/h. Estimates of the repair parameters using different types of analysis revealed an α/β ratio of 2.44 Gy and a (mono- exponential) half time of repair (=t (1(2)) ) of 1.43 hours; for the maximum of 150 % of the prescribed dose these values were 3.67 Gy and 1.43 hours respectively. Conclusions: Spinal cord radiation tolerance is

  20. Radiobiological equivalent of low/high dose rate brachytherapy and evaluation of tumor and normal responses to the dose.

    Science.gov (United States)

    Manimaran, S

    2007-06-01

    The aim of this study was to compare the biological equivalent of low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy in terms of the more recent linear quadratic (LQ) model, which leads to theoretical estimation of biological equivalence. One of the key features of the LQ model is that it allows a more systematic radiobiological comparison between different types of treatment because the main parameters alpha/beta and micro are tissue-specific. Such comparisons also allow assessment of the likely change in the therapeutic ratio when switching between LDR and HDR treatments. The main application of LQ methodology, which focuses on by increasing the availability of remote afterloading units, has been to design fractionated HDR treatments that can replace existing LDR techniques. In this study, with LDR treatments (39 Gy in 48 h) equivalent to 11 fractions of HDR irradiation at the experimental level, there are increasing reports of reproducible animal models that may be used to investigate the biological basis of brachytherapy and to help confirm theoretical predictions. This is a timely development owing to the nonavailability of sufficient retrospective patient data analysis. It appears that HDR brachytherapy is likely to be a viable alternative to LDR only if it is delivered without a prohibitively large number of fractions (e.g., fewer than 11). With increased scientific understanding and technological capability, the prospect of a dose equivalent to HDR brachytherapy will allow greater utilization of the concepts discussed in this article.

  1. Radiobiological modelling of dose-gradient effects in low dose rate, high dose rate and pulsed brachytherapy

    International Nuclear Information System (INIS)

    Armpilia, C; Dale, R G; Sandilos, P; Vlachos, L

    2006-01-01

    This paper presents a generalization of a previously published methodology which quantified the radiobiological consequences of dose-gradient effects in brachytherapy applications. The methodology uses the linear-quadratic (LQ) formulation to identify an equivalent biologically effective dose (BED eq ) which, if applied uniformly to a specified tissue volume, would produce the same net cell survival as that achieved by a given non-uniform brachytherapy application. Multiplying factors (MFs), which enable the equivalent BED for an enclosed volume to be estimated from the BED calculated at the dose reference surface, have been calculated and tabulated for both spherical and cylindrical geometries. The main types of brachytherapy (high dose rate (HDR), low dose rate (LDR) and pulsed (PB)) have been examined for a range of radiobiological parameters/dimensions. Equivalent BEDs are consistently higher than the BEDs calculated at the reference surface by an amount which depends on the treatment prescription (magnitude of the prescribed dose) at the reference point. MFs are closely related to the numerical BED values, irrespective of how the original BED was attained (e.g., via HDR, LDR or PB). Thus, an average MF can be used for a given prescribed BED as it will be largely independent of the assumed radiobiological parameters (radiosensitivity and α/β) and standardized look-up tables may be applicable to all types of brachytherapy treatment. This analysis opens the way to more systematic approaches for correlating physical and biological effects in several types of brachytherapy and for the improved quantitative assessment and ranking of clinical treatments which involve a brachytherapy component

  2. Dosimetric evaluation of PLATO and Oncentra treatment planning systems for High Dose Rate (HDR) brachytherapy gynecological treatments

    Energy Technology Data Exchange (ETDEWEB)

    Singh, Hardev; De La Fuente Herman, Tania; Showalter, Barry; Thompson, Spencer J.; Syzek, Elizabeth J.; Herman, Terence; Ahmad, Salahuddin [Department of Radiation Oncology, Peggy and Charles Stephenson Oklahoma Cancer Center, University of Oklahoma Health Sciences Center, Oklahoma City, OK 73104 (United States)

    2012-10-23

    This study compares the dosimetric differences in HDR brachytherapy treatment plans calculated with Nucletron's PLATO and Oncentra MasterPlan treatment planning systems (TPS). Ten patients (1 T1b, 1 T2a, 6 T2b, 2 T4) having cervical carcinoma, median age of 43.5 years (range, 34-79 years) treated with tandem and ring applicator in our institution were selected retrospectively for this study. For both Plato and Oncentra TPS, the same orthogonal films anterior-posterior (AP) and lateral were used to manually draw the prescription and anatomical points using definitions from the Manchester system and recommendations from the ICRU report 38. Data input for PLATO was done using a digitizer and Epson Expression 10000XL scanner was used for Oncentra where the points were selected on the images in the screen. The prescription doses for these patients were 30 Gy to points right A (RA) and left A (LA) delivered in 5 fractions with Ir-192 HDR source. Two arrangements: one dwell position and two dwell positions on the tandem were used for dose calculation. The doses to the patient points right B (RB) and left B (LB), and to the organs at risk (OAR), bladder and rectum for each patient were calculated. The mean dose and the mean percentage difference in dose calculated by the two treatment planning systems were compared. Paired t-tests were used for statistical analysis. No significant differences in mean RB, LB, bladder and rectum doses were found with p-values > 0.14. The mean percent difference of doses in RB, LB, bladder and rectum are found to be less than 2.2%, 1.8%, 1.3% and 2.2%, respectively. Dose calculations based on the two different treatment planning systems were found to be consistent and the treatment plans can be made with either system in our department without any concern.

  3. High-dose-rate intracavitary brachytherapy (HDR-IC) in treatment of cervical carcinoma: 5-year results and implication of increased low-grade rectal complication on initiation of an HDR-IC fractionation scheme

    International Nuclear Information System (INIS)

    Wang Chongjong; Wan Leung, Stephen; Chen Huichun; Sun Limin; Fang Fumin; Changchien Chanchao; Huang Engyen; Wu Jiaming; Chen Chuhnchih

    1997-01-01

    Purpose: To report the treatment results and rectal/bladder complications of cervical carcinoma radically treated with high-dose-rate intracavitary brachytherapy (HDR-IC). The current policy of using three-fraction scheme was examined. Methods and Materials: Between November 1987 and August 1990, 173 patients with cervical carcinoma were treated with curative-intent radiation therapy. Whole pelvic irradiation was administered with 10-MV X ray. Dose to the central cervix was 40-44 Gy in 20-22 fractions, following by pelvic wall boost 6-14 Gy in three to seven fractions with central shielding. 60 Co sources were used for HDR-IC, and 7.2 Gy was given to Point A for three applications, 1-2 weeks apart. Duration of follow-up was 5-7.8 years. Results: Twenty-eight patients (16%) developed central-regional recurrences. Overall 5-year actuarial pelvic control rate was 83%. By stage, 5-year actuarial pelvic control rates were 94%, 87%, and 72% for Stages IB + IIA, IIB + IIIA, and IIIB + IVA, respectively. Thirty-one patients (18%) developed distant metastasis. Overall 5-year actuarial survival rate was 58%. By stage, 5-year actuarial survival rates were 79%, 59%, and 41% for Stages IB + IIA, IIB + IIIA, and IIIB + IVA, respectively. Sixty-six (38%) and 19 patients (11%) developed rectal and bladder complications, respectively. For rectal complication, the overall actuarial rate was 38% at 5 years. By grade, 5-year actuarial rectal complication rates were 24%, 15%, 4%, and 3% for Grades 1-4, respectively. Overall prevalence of rectal complications was 37% and 14% at 2 and 5 years, respectively. Prevalence of low-grade rectal complication (Grades 1 and 2) was dominant at 2 years (30%), but declined to 8% at 5 years. Prevalence of high-grade, severe rectal complication (Grades 3 and 4) remained steady at 2 and 5 years (7% and 6%, respectively). Five-year actuarial bladder complication was 9%. Five-year prevalence of bladder complication was 2%. Conclusion: Using a three

  4. Phase III trial of high- vs. low-dose-rate interstitial radiotherapy for early mobile tongue cancer

    International Nuclear Information System (INIS)

    Inoue, Takehiro; Inoue, Toshihiko; Yoshida, Ken; Yoshioka, Yasuo; Shimamoto, Shigetoshi; Tanaka, Eiichi; Yamazaki, Hideya; Shimizutani, Kimishige; Teshima, Teruki; Furukawa, Souhei

    2001-01-01

    Purpose: Early mobile tongue cancer can be controlled with interstitial radiotherapy (ISRT). We carried out a Phase III trial to compare the treatment results of low-dose-rate (Ld) ISRT and high-dose-rate (HDR) ISRT for early mobile tongue cancer. Methods and Materials: From April 1992 through October 1996, 59 patients with cancer of the early mobile tongue were registered in this Phase III study. Eight patients were excluded from the evaluation because of violations of the requirements for this study. Of 51 eligible patients, 26 patients were treated with LDR-ISRT (70 Gy/4-9 days) and 25 patients with HDR-ISRT (60 Gy/10 fractions/1 week). For the hyperfractionated HDR-ISRT, the time interval between 2 fractions was more than 6 h. Results: Five-year local control rates of the LDR and HDR groups were 84% and 87% respectively. Nodal metastasis occurred in 6 patients in each group. Five-year nodal control rates of the LDR and HDR groups were 77% and 76%, respectively. Conclusion: Hyperfractionated HDR-ISRT for early mobile tongue cancer has the same local control compared with continuous LDR-ISRT. Hyperfractionated HDR-ISRT is an alternative treatment for continuous LDR-ISRT

  5. Biological effective doses in the intracavitary high dose rate brachytherapy of cervical cancer

    Directory of Open Access Journals (Sweden)

    Y. Sobita Devi

    2011-12-01

    Full Text Available Purpose: The aim of this study is to evaluate the decrease of biological equivalent dose and its correlation withlocal/loco-regional control of tumour in the treatment of cervical cancer when the strength of the Ir-192 high dose rate(HDR brachytherapy (BT source is reduced to single, double and triple half life in relation to original strength of10 Ci (~ 4.081 cGy x m2 x h–1. Material and methods: A retrospective study was carried out on 52 cervical cancer patients with stage II and IIItreated with fractionated HDR-BT following external beam radiation therapy (EBRT. International Commission onRadiation Units and Measurement (ICRU points were defined according to ICRU Report 38, using two orthogonal radiographimages taken by Simulator (Simulix HQ. Biologically effective dose (BED was calculated at point A for diffe -rent Ir-192 source strength and its possible correlation with local/loco-regional tumour control was discussed. Result: The increase of treatment time per fraction of dose due to the fall of dose rate especially in HDR-BT of cervicalcancer results in reduction in BED of 2.59%, 7.02% and 13.68% with single, double and triple half life reduction ofsource strength, respectively. The probabilities of disease recurrence (local/loco-regional within 26 months are expectedas 0.12, 0.12, 0.16, 0.39 and 0.80 for source strength of 4.081, 2.041, 1.020, 0.510 and 0.347 cGy x m2 x h–1, respectively.The percentages of dose increase required to maintain the same BED with respect to initial BED were estimated as1.71, 5.00, 11.00 and 15.86 for the dose rate of 24.7, 12.4, 6.2 and 4.2 Gy/hr at point A, respectively. Conclusions: This retrospective study of cervical cancer patients treated with HDR-BT at different Ir-192 sourcestrength shows reduction in disease free survival according to the increase in treatment time duration per fraction.The probable result could be associated with the decrease of biological equivalent dose to point A. Clinical

  6. High versus low dose rate intracavitary irradiation for adenocarcinoma of the uterine cervix

    International Nuclear Information System (INIS)

    Kim, Woo-Chul; Loh, John J.K.; Kim, Gwi-Eon; Suh, Chang-Ok

    2001-01-01

    Traditionally, low dose rate (LDR) brachytherapy has been used as a standard modality in the treatment of patients with carcinoma of the uterine cervix. The purpose of this work was to evaluate the effects of high dose rate (HDR) brachytherapy on patients with adenocarcinoma of the uterine cervix and to compare them with the effects of LDR brachytherapy. From January 1971 to December 1992, 104 patients suffering from adenocarcinoma of the uterine cervix were treated with radiation therapy in the Department of Radiation Oncology, Yonsei University. LDR brachytherapy was carried out on 34 patients and HDR brachytherapy on 70 patients. In the LDR group, eight patients were in stage IB, six in IIA, 12 in IIB, three in IIIA and five in IIIB. External radiation therapy was delivered with 10 MV X-rays, 2 Gy fraction per day, total dose of whole pelvis 36-52 Gy (median 46 Gy). LDR radium intracavitary irradiation was performed with a Henschke applicator, 37-59 Gy targeted at point A (median 43 Gy). In the HDR group, there were 16 patients in stage IB, six in IIA, 32 in IIB and 16 in IIIB. The total whole pelvis dose of external radiation was 40-50 Gy (median 44 Gy), daily 1.8-2.0 Gy. HDR Co-60 intracavitary irradiation was performed with a remotely controlled after-loading system (RALS), 30-48 Gy (median 39 Gy) targeted at point A, three times per week, 3 Gy per fraction. The 5-year overall survival rate in the LDR group was 72.9, 61.9 and 35.7% in stage I, II and III, respectively and the corresponding figures for HDR were 87.1, 58.3 and 43.8% (p 0.05). No prognostic factors were evident in the comparison between the two groups. There was no difference in terms of 5-year survival rate in the patients with adenocarcinoma of the uterine cervix between those treated with HDR and those treated with LDR brachytherapy. Even though late complication rates were higher in the HDR group, most of them were classified as grade I. This retrospective study suggests that HDR

  7. High dose rate brachytherapy for the palliation of malignant dysphagia

    International Nuclear Information System (INIS)

    Homs, Marjolein Y.V.; Eijkenboom, Wilhelmina M.H.; Coen, Veronique L.M.A.; Haringsma, Jelle; Blankenstein, Mark van; Kuipers, Ernst J.; Siersema, Peter D.

    2003-01-01

    Background and purpose: High dose rate (HDR) brachytherapy is a commonly used palliative treatment for esophageal carcinoma. We evaluated the outcome of HDR brachytherapy in patients with malignant dysphagia. Material and methods: A retrospective analysis over a 10-year period was performed of 149 patients treated with HDR brachytherapy, administered in one or two sessions, at a median dose of 15 Gy. Patients were evaluated for functional outcome, complications, recurrent dysphagia, and survival. Results: At 6 weeks after HDR brachytherapy, dysphagia scores had improved from a median of 3 to 2 (n=104; P<0.001), however, dysphagia had not improved in 51 (49%) patients. Procedure-related complications occurred in seven (5%) patients. Late complications, including fistula formation or bleeding, occurred in 11 (7%) patients. Twelve (8%) patients experienced minor retrosternal pain. Median survival of the patients was 160 days with a 1-year survival rate of 15%. Procedure-related mortality was 2%. At follow-up, 55 (37%) patients experienced recurrent dysphagia. In 34 (23%) patients a metal stent was placed to relieve persistent or recurrent dysphagia. Conclusion: HDR brachytherapy is a moderately effective treatment for the palliation of malignant dysphagia. The incidence of early major complications is low, however, persistent and recurrent dysphagia occur frequently, and require often additional treatment

  8. Anatomy-based inverse planning dose optimization in HDR prostate implant: A toxicity study

    International Nuclear Information System (INIS)

    Mahmoudieh, Alireza; Tremblay, Christine; Beaulieu, Luc; Lachance, Bernard; Harel, Francois; Lessard, Etienne; Pouliot, Jean; Vigneault, Eric

    2005-01-01

    Background and purpose: The aim of this study is to evaluate the acute and late complications in patients who have received HDR implant boost using inverse planning, and to determine dose volume correlations. Patients and methods: Between September 1999 and October 2002, 44 patients with locally advanced prostate cancer (PSA ≥10 ng/ml, and/or Gleason score ≥7, and/or Stage T2c or higher) were treated with 40-45 Gy external pelvic field followed by 2-3 fraction of inverse-planned HDR implant boost (6-9.5 Gy /fraction). Median follow-up time was 1.7 years with 81.8% of patients who had at least 12 months of follow up (range 8.6-42.5. Acute and late morbidity data were collected and graded according to RTOG criteria. Questionnaires were used to collect prostate related measures of quality of life, and international prostate symptom score (IPSS) before and after treatment. Dose-volume histograms for prostate, urethra, bladder, penis bulb and rectum were analyzed. Results: The median patient age was 64 years. Of these, 32% were in the high risk group, and 61% in the intermediate risk group. 3 patients (7%) had no adverse prognostic factors. A single grade 3 GU acute toxicity was reported but no grade 3-4 acute GI toxicity. No grade 3-4 late GU or GI toxicity was reported. Acute (late) grade 2 urinary and rectal symptoms were reported in 31.8 (11.4%) and 4.6% (4.6%) of patients, respectively. A trend for predicting acute GU toxicity is seen for total HDR dose of more than 18 Gy (OR=3.6, 95%CI=[0.96-13.5], P=0.058). The evolution of toxicity is presented for acute and late GU/GI toxicity. Erectile dysfunction occurs in approximately 27% of patients who were not on hormonal deprivation, but may be taking sildenafil. The IPSS peaked on averaged 6 weeks post-implant and returned to the baseline at a median of 6 months. Conclusions: Inverse-planned HDR brachytherapy is a viable option to deliver higher dose to the prostate as a boost without increasing GU or rectal

  9. MRI dosimetry using an echo-quotient technique for high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Ansbacher, W.

    1996-01-01

    MRI gel dosimetry is a relatively new technique that has many advantages over conventional methods, and is particularly suited to High Dose Rate (HDR) Brachytherapy. The dosimeter has high spatial resolution and a water-equivalent response over a wide range of photon energies. Because it is an integrating dosimeter, it allows for efficient mapping of the dynamically-produced distributions from an HDR source. As an example of this technique, the dose response, which is calibrated in terms of the change in spin-spin relaxation time, has been used to investigate the anisotropy of an HDR source. (author). 1 fig

  10. Dose escalation using conformal high-dose-rate brachytherapy improves outcome in unfavorable prostate cancer

    International Nuclear Information System (INIS)

    Martinez, Alvaro A.; Gustafson, Gary; Gonzalez, Jose; Armour, Elwood; Mitchell, Chris; Edmundson, Gregory; Spencer, William; Stromberg, Jannifer; Huang, Raywin; Vicini, Frank

    2002-01-01

    Purpose: To overcome radioresistance for patients with unfavorable prostate cancer, a prospective trial of pelvic external beam irradiation (EBRT) interdigitated with dose-escalating conformal high-dose-rate (HDR) prostate brachytherapy was performed. Methods and Materials: Between November 1991 and August 2000, 207 patients were treated with 46 Gy pelvic EBRT and increasing HDR brachytherapy boost doses (5.50-11.5 Gy/fraction) during 5 weeks. The eligibility criteria were pretreatment prostate-specific antigen level ≥10.0 ng/mL, Gleason score ≥7, or clinical Stage T2b or higher. Patients were divided into 2 dose levels, low-dose biologically effective dose 93 Gy (149 patients). No patient received hormones. We used the American Society for Therapeutic Radiology and Oncology definition for biochemical failure. Results: The median age was 69 years. The mean follow-up for the group was 4.4 years, and for the low and high-dose levels, it was 7.0 and 3.4 years, respectively. The actuarial 5-year biochemical control rate was 74%, and the overall, cause-specific, and disease-free survival rate was 92%, 98%, and 68%, respectively. The 5-year biochemical control rate for the low-dose group was 52%; the rate for the high-dose group was 87% (p<0.001). Improvement occurred in the cause-specific survival in favor of the brachytherapy high-dose level (p=0.014). On multivariate analysis, a low-dose level, higher Gleason score, and higher nadir value were associated with increased biochemical failure. The Radiation Therapy Oncology Group Grade 3 gastrointestinal/genitourinary complications ranged from 0.5% to 9%. The actuarial 5-year impotency rate was 51%. Conclusion: Pelvic EBRT interdigitated with transrectal ultrasound-guided real-time conformal HDR prostate brachytherapy boost is both a precise dose delivery system and a very effective treatment for unfavorable prostate cancer. We demonstrated an incremental beneficial effect on biochemical control and cause

  11. LDR vs. HDR brachytherapy for localized prostate cancer: the view from radiobiological models.

    Science.gov (United States)

    King, Christopher R

    2002-01-01

    Permanent LDR brachytherapy and temporary HDR brachytherapy are competitive techniques for clinically localized prostate radiotherapy. Although a randomized trial will likely never be conducted comparing these two forms of brachytherapy, a comparative radiobiological modeling analysis proves useful in understanding some of their intrinsic differences, several of which could be exploited to improve outcomes. Radiobiological models based upon the linear quadratic equations are presented for fractionated external beam, fractionated (192)Ir HDR brachytherapy, and (125)I and (103)Pd LDR brachytherapy. These models incorporate the dose heterogeneities present in brachytherapy based upon patient-derived dose volume histograms (DVH) as well as tumor doubling times and repair kinetics. Radiobiological parameters are normalized to correspond to three accepted clinical risk factors based upon T-stage, PSA, and Gleason score to compare models with clinical series. Tumor control probabilities (TCP) for LDR and HDR brachytherapy (as monotherapy or combined with external beam) are compared with clinical bNED survival rates. Predictions are made for dose escalation with HDR brachytherapy regimens. Model predictions for dose escalation with external beam agree with clinical data and validate the models and their underlying assumptions. Both LDR and HDR brachytherapy achieve superior tumor control when compared with external beam at conventional doses (LDR brachytherapy as boost achieves superior tumor control than when used as monotherapy. Stage for stage, both LDR and current HDR regimens achieve similar tumor control rates, in agreement with current clinical data. HDR monotherapy with large-dose fraction sizes might achieve superior tumor control compared with LDR, especially if prostate cancer possesses a high sensitivity to dose fractionation (i.e., if the alpha/beta ratio is low). Radiobiological models support the current clinical evidence for equivalent outcomes in localized

  12. Application of the Monte Carlo integration method in calculations of dose distributions in HDR-Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Baltas, D; Geramani, K N; Ioannidis, G T; Kolotas, C; Zamboglou, N [Strahlenklinik, Stadtische Kliniken Offenbach, Offenbach (Germany); Giannouli, S [Department of Electrical and Computer Engineering, National Technical University of Athens, Athens (Greece)

    1999-12-31

    Source anisotropy is a very important factor in brachytherapy quality assurance of high dose rate HDR Ir 192 afterloading stepping sources. If anisotropy is not taken into account then doses received by a brachytherapy patient in certain directions can be in error by a clinically significant amount. Experimental measurements of anisotropy are very labour intensive. We have shown that within acceptable limits of accuracy, Monte Carlo integration (MCI) of a modified Sievert integral (3D generalisation) can provide the necessary data within a much shorter time scale than can experiments. Hence MCI can be used for routine quality assurance schedules whenever a new design of HDR or PDR Ir 192 is used for brachytherapy afterloading. Our MCI calculation results are comparable with published experimental data and Monte Carlo simulation data for microSelectron and VariSource Ir 192 sources. We have shown not only that MCI offers advantages over alternative numerical integration methods, but also that treating filtration coefficients as radial distance-dependent functions improves Sievert integral accuracy at low energies. This paper also provides anisotropy data for three new Ir 192 sources, one for microSelectron-HDR and two for the microSelectron-PDR, for which data currently is not available. The information we have obtained in this study can be incorporated into clinical practice.

  13. Remote Afterloading High Dose Rate (HDR) Endobronchial Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Chang, Hyesook; Choi, Eun Kyung; Yi, Byong Yong; Kim, Won Dong; Kim, Woo Sung; Koh, Youn Suck [Ulsan University College of Medicine, Seoul (Korea, Republic of)

    1991-12-15

    Authors described the remote afterloading endobronchial brachytherapy (EBBT) technique using the microSelectron HDR Ir-192 and the Asan Medical Center experience. Total 28 EBBT in 9 patients were performed since November 1989 and 24 EBBT in 8 patients were employed for palliation and 3 EBBT in 1 patient was treated curatively. Authors observed a significant relief of obstructive symptom with tumor regression in 7 patients out of 8 who were treated palliatively but one of them died of pulmonary congestion in 3 weeks after EBBT. One patient with prior therapy of extensive electrocautery expired within 1 day after 2nd EBBT procedure with massive hemorrhage from the lesion. EBBT procedure has been tolerable and can be performed as an outpatient.

  14. Remote Afterloading High Dose Rate (HDR) Endobronchial Brachytherapy

    International Nuclear Information System (INIS)

    Chang, Hyesook; Choi, Eun Kyung; Yi, Byong Yong; Kim, Won Dong; Kim, Woo Sung; Koh, Youn Suck

    1991-01-01

    Authors described the remote afterloading endobronchial brachytherapy (EBBT) technique using the microSelectron HDR Ir-192 and the Asan Medical Center experience. Total 28 EBBT in 9 patients were performed since November 1989 and 24 EBBT in 8 patients were employed for palliation and 3 EBBT in 1 patient was treated curatively. Authors observed a significant relief of obstructive symptom with tumor regression in 7 patients out of 8 who were treated palliatively but one of them died of pulmonary congestion in 3 weeks after EBBT. One patient with prior therapy of extensive electrocautery expired within 1 day after 2nd EBBT procedure with massive hemorrhage from the lesion. EBBT procedure has been tolerable and can be performed as an outpatient

  15. Audits in high dose rate brachytherapy in Brazil

    International Nuclear Information System (INIS)

    Marechal, M.H.; Rosa, L.A.; Velasco, A.; Paiva, E. de; Goncalves, M.; Castelo, L.C.

    2002-01-01

    The lack of well established dosimetry protocols for HDR sources is a point of great concern regarding the uniformity of procedures within a particular country. The main objective of this paper is to report the results of an implementation of the audit program in dosimetry of high dose rate brachytherapy sources used by the radiation therapy centers in Brazil. In Brazil, among 169 radiotherapy centers, 35 have HDR brachytherapy systems. This program started in August 2001 and until now eight radiotherapy services were audited. The audit program consists of the visit in loco to each center and the evaluation of the intensity of the source with a well type chamber specially design for HDR 192 Ir sources. The measurements was carried out with a HDR1000PLUS Brachytherapy Well Type Chamber and a MAX 4000 Electrometer, both manufactured by Standard Imaging Inc. The chamber was calibrated in air kerma strength by the Accredited Dosimetry Calibration Laboratory, Department of Medical Physics, University of Wisconsin in the USA. The same chamber was calibrated in Brazil using a 192 lr high dose rate source whose intensity was determined by 60 Co gamma rays and 250 kV x rays interpolation methodology. The Nk of 60 Co and 250 kV x rays were provided by the Brazilian National Standard Laboratory for Ionizing Radiation (LMNRI)

  16. SU-G-201-01: An Automated Treatment Plan Quality Assurance Program for High-Dose Rate (HDR) Brachytherapy with a VaginalCylinder Applicator

    Energy Technology Data Exchange (ETDEWEB)

    Zhou, Y; Tan, J; Jiang, S; Albuquerque, K; Jia, X [UT Southwestern Medical Center, Dallas, TX (United States)

    2016-06-15

    Purpose: Plan specific quality assurance (QA) is an important step in high dose rate (HDR) brachytherapy to ensure the integrity of a treatment plan. The conventional approach is to assemble a set of plan screen-captures in a document and have an independent plan-checker to verify it. Not only is this approach cumbersome and time-consuming, using a document also limits the items that can be verified, hindering plan quality and patient safety. We have initiated efforts to develop a web-based HDR brachytherapy QA system called AutoBrachy QA, for comprehensive and efficient QA. This abstract reports a new plugin in this system for the QA of a cylinder HDR brachytherapy treatment. Methods: A cylinder plan QA module was developed using Python. It was plugged into our AutoBrachy QA system. This module extracted information from CT images and treatment plan. Image processing techniques were employed to obtain geometric parameters, e.g. cylinder diameter. A comprehensive set of eight geometrical and eight dosimetric features of the plan were validated against user specified planning parameter, such as prescription value, treatment depth and length, etc. A PDF document was generated, consisting of a summary QA sheet with all the QA results, as well as images showing plan details. Results: The cylinder QA program has been implemented in our clinic. To date, it has been used in 11 patient cases and was able to successfully perform QA tests in all of them. The QA program reduced the average plan QA time from 7 min using conventional manual approach to 0.5 min. Conclusion: Being a new module in our AutoBrachy QA system, an automated treatment plan QA module for cylinder HDR brachytherapy has been successfully developed and clinically implemented. This module improved clinical workflow and plan integrity compared to the conventional manual approach.

  17. SU-G-201-01: An Automated Treatment Plan Quality Assurance Program for High-Dose Rate (HDR) Brachytherapy with a VaginalCylinder Applicator

    International Nuclear Information System (INIS)

    Zhou, Y; Tan, J; Jiang, S; Albuquerque, K; Jia, X

    2016-01-01

    Purpose: Plan specific quality assurance (QA) is an important step in high dose rate (HDR) brachytherapy to ensure the integrity of a treatment plan. The conventional approach is to assemble a set of plan screen-captures in a document and have an independent plan-checker to verify it. Not only is this approach cumbersome and time-consuming, using a document also limits the items that can be verified, hindering plan quality and patient safety. We have initiated efforts to develop a web-based HDR brachytherapy QA system called AutoBrachy QA, for comprehensive and efficient QA. This abstract reports a new plugin in this system for the QA of a cylinder HDR brachytherapy treatment. Methods: A cylinder plan QA module was developed using Python. It was plugged into our AutoBrachy QA system. This module extracted information from CT images and treatment plan. Image processing techniques were employed to obtain geometric parameters, e.g. cylinder diameter. A comprehensive set of eight geometrical and eight dosimetric features of the plan were validated against user specified planning parameter, such as prescription value, treatment depth and length, etc. A PDF document was generated, consisting of a summary QA sheet with all the QA results, as well as images showing plan details. Results: The cylinder QA program has been implemented in our clinic. To date, it has been used in 11 patient cases and was able to successfully perform QA tests in all of them. The QA program reduced the average plan QA time from 7 min using conventional manual approach to 0.5 min. Conclusion: Being a new module in our AutoBrachy QA system, an automated treatment plan QA module for cylinder HDR brachytherapy has been successfully developed and clinically implemented. This module improved clinical workflow and plan integrity compared to the conventional manual approach.

  18. Results of radiation therapy for uterine cervical cancer using high dose rate remote after loading system

    International Nuclear Information System (INIS)

    Ogawa, Yoshihiro; Nemoto, Kenji

    2003-01-01

    In Japan, radiotherapy with high dose rate remote after loading system (HDR-RALS) for intracavitary brachytherapy is the standard treatment for more than 30 years. This report showed the usefulness of HDR-RALS for uterine cervical cancer. From 1980 through 1999, 442 patients with uterine cervical cancers (stage I: 66, stage II: 161, stage III: 165, stage IV: 50) were treated. Radiotherapy was performed both external teletherapy and HDR-RALS. Overall survival rate at 5 years was 60.2%. The 5-year actuarial incidence of all complications was 16.4%. The 5-year actuarial incidence of all complications in cases treated with the sum doses of whole pelvic irradiation (without central shield) and RALS up to 49 Gy, 50 to 59 Gy or larger doses were 7.5%, 11.0% and 25.2%, respectively. Radiation therapy using HDR-RALS was very effective. While the dose of whole pelvic irradiation was increased, the actuarial incidence of all complications was increased. (author)

  19. Optimized dose distribution of a high dose rate vaginal cylinder

    International Nuclear Information System (INIS)

    Li Zuofeng; Liu, Chihray; Palta, Jatinder R.

    1998-01-01

    Purpose: To present a comparison of optimized dose distributions for a set of high-dose-rate (HDR) vaginal cylinders calculated by a commercial treatment-planning system with benchmark calculations using Monte-Carlo-calculated dosimetry data. Methods and Materials: Optimized dose distributions using both an isotropic and an anisotropic dose calculation model were obtained for a set of HDR vaginal cylinders. Mathematical optimization techniques available in the computer treatment-planning system were used to calculate dwell times and positions. These dose distributions were compared with benchmark calculations with TG43 formalism and using Monte-Carlo-calculated data. The same dwell times and positions were used for a quantitative comparison of dose calculated with three dose models. Results: The isotropic dose calculation model can result in discrepancies as high as 50%. The anisotropic dose calculation model compared better with benchmark calculations. The differences were more significant at the apex of the vaginal cylinder, which is typically used as the prescription point. Conclusion: Dose calculation models available in a computer treatment-planning system must be evaluated carefully to ensure their correct application. It should also be noted that when optimized dose distribution at a distance from the cylinder surface is calculated using an accurate dose calculation model, the vaginal mucosa dose becomes significantly higher, and therefore should be carefully monitored

  20. A dosimetric selectivity intercomparison of HDR brachytherapy, IMRT and helical tomotherapy in prostate cancer radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Hermesse, Johanne; Biver, Sylvie; Jansen, Nicolas; Coucke, Philippe [Dept. of Radiation Oncology, Liege Univ. Hospital (Belgium); Lenaerts, Eric [Dept. of Medical Physics, Liege Univ. Hospital (Belgium); De Patoul, Nathalie; Vynckier, Stefaan [Dept. of Medical Physics, St Luc Univ. Hospital, Brussels (Belgium); Scalliet, Pierre [Dept. of Radiation Oncology, St Luc Univ. Hospital, Brussels (Belgium); Nickers, Philippe [Dept. of Radiation Oncology, Oscar Lambret Center, Lille (France)

    2009-11-15

    Background and purpose: dose escalation in order to improve the biochemical control in prostate cancer requires the application of irradiation techniques with high conformality. The dosimetric selectivity of three radiation modalities is compared: high-dose-rate brachytherapy (HDR-BT), intensity-modulated radiation radiotherapy (IMRT), and helical tomotherapy (HT). Patients and methods: ten patients with prostate adenocarcinoma treated by a 10-Gy HDR-BT boost after external-beam radiotherapy were investigated. For each patient, HDR-BT, IMRT and HT theoretical treatment plans were realized using common contour sets. A 10-Gy dose was prescribed to the planning target volume (PTV). The PTVs and critical organs' dose-volume histograms obtained were compared using Student's t-test. Results: HDR-BT delivers spontaneously higher mean doses to the PTV with smaller cold spots compared to IMRT and HT. 33% of the rectal volume received a mean HDR-BT dose of 3.86 {+-} 0.3 Gy in comparison with a mean IMRT dose of 6.57 {+-} 0.68 Gy and a mean HT dose of 5.58 {+-} 0.71 Gy (p < 0.0001). HDR-BT also enables to better spare the bladder. The hot spots inside the urethra are greater with HDR-BT. The volume of healthy tissue receiving 10% of the prescribed dose is reduced at least by a factor of 8 with HDR-BT (p < 0.0001). Conclusion: HDR-BT offers better conformality in comparison with HT and IMRT and reduces the volume of healthy tissue receiving a low dose. (orig.)

  1. Salvage high-dose-rate brachytherapy for local prostate cancer recurrence after radical radiotherapy

    Directory of Open Access Journals (Sweden)

    V. A. Solodkiy

    2016-01-01

    Full Text Available Studies salvage interstitial radiation therapy for recurrent prostate cancer, launched at the end of the XX century. In recent years, more and more attention is paid to high-dose-rate brachytherapy (HDR-BT as a method of treating local recurrence.The purpose of research – preliminary clinical results of salvage high-dose-rate brachytherapy applied in cases of suspected local recurrence or of residual tumour after radiotherapy.Preliminary findings indicate the possibility of using HDR-BT, achieving local tumor control with low genitourinary toxicity.

  2. Gamma Low-Dose-Rate Ionizing Radiation Stimulates Adaptive Functional and Molecular Response in Human Aortic Endothelial Cells in a Threshold-, Dose-, and Dose Rate–Dependent Manner

    Science.gov (United States)

    Vieira Dias, Juliana; Gloaguen, Celine; Kereselidze, Dimitri; Manens, Line; Tack, Karine; Ebrahimian, Teni G

    2018-01-01

    A central question in radiation protection research is whether low-dose and low-dose-rate (LDR) exposures to ionizing radiation play a role in progression of cardiovascular disease. The response of endothelial cells to different LDR exposures may help estimate risk of cardiovascular disease by providing the biological mechanism involved. We investigated the effect of chronic LDR radiation on functional and molecular responses of human aorta endothelial cells (HAoECs). Human aorta endothelial cells were continuously irradiated at LDR (6 mGy/h) for 15 days and analyzed at time points when the cumulative dose reached 0.05, 0.5, 1.0, and 2.0 Gy. The same doses were administered acutely at high-dose rate (HDR; 1 Gy/min). The threshold for the loss of angiogenic capacity for both LDR and HDR radiations was between 0.5 and 1.0 Gy. At 2.0 Gy, angiogenic capacity returned to normal only for HAoEC exposed to LDR radiation, associated with increased expression of antioxidant and anti-inflammatory genes. Pre-LDR, but not pre-HDR, radiation, followed by a single acute 2.0 Gy challenge dose sustained the expression of antioxidant and anti-inflammatory genes and stimulated angiogenesis. Our results suggest that dose rate is important in cellular response and that a radioadaptive response is involved for a 2.0 Gy dose at LDR. PMID:29531508

  3. Biological effective dose evaluation in gynaecological brachytherapy: LDR and HDR treatments, dependence on radiobiological parameters, and treatment optimisation.

    Science.gov (United States)

    Bianchi, C; Botta, F; Conte, L; Vanoli, P; Cerizza, L

    2008-10-01

    This study was undertaken to compare the biological efficacy of different high-dose-rate (HDR) and low-dose-rate (LDR) treatments of gynaecological lesions, to identify the causes of possible nonuniformity and to optimise treatment through customised calculation. The study considered 110 patients treated between 2001 and 2006 with external beam radiation therapy and/or brachytherapy with either LDR (afterloader Selectron, (137)Cs) or HDR (afterloader microSelectron Classic, (192)Ir). The treatments were compared in terms of biologically effective dose (BED) to the tumour and to the rectum (linear-quadratic model) by using statistical tests for comparisons between independent samples. The difference between the two treatments was statistically significant in one case only. However, within each technique, we identified considerable nonuniformity in therapeutic efficacy due to differences in fractionation schemes and overall treatment time. To solve this problem, we created a Microsoft Excel spreadsheet allowing calculation of the optimal treatment for each patient: best efficacy (BED(tumour)) without exceeding toxicity threshold (BED(rectum)). The efficacy of a treatment may vary as a result of several factors. Customised radiobiological evaluation is a useful adjunct to clinical evaluation in planning equivalent treatments that satisfy all dosimetric constraints.

  4. Australian high-dose-rate brachytherapy protocols for gynaecological malignancy

    International Nuclear Information System (INIS)

    MacLeod, C.; Dally, M.; Stevens, M.; Thornton, D.; Carruthers, S.; Jeal, P.

    2001-01-01

    There is no consensus over the optimal dose fractionation schedules for high-dose-rate (HDR) brachytherapy used for gynaecological malignancy. In Australian public hospital departments of radiation oncology, HDR brachytherapy for gynaecological cancer is being more commonly used. A survey of public departments that are using this technology, or that plan to introduce this technology, was performed. Their current protocols are presented. In general, protocols are similar biologically; however, the practical aspects such as the number of fractions given do vary and may reflect resource restrictions or, alternatively, differences in interpretations of the literature and of the best protocols by clinicians. Copyright (2001) Blackwell Science Pty Ltd

  5. High versus low-dose rate brachytherapy for cervical cancer.

    Science.gov (United States)

    Patankar, Sonali S; Tergas, Ana I; Deutsch, Israel; Burke, William M; Hou, June Y; Ananth, Cande V; Huang, Yongmei; Neugut, Alfred I; Hershman, Dawn L; Wright, Jason D

    2015-03-01

    Brachytherapy plays an important role in the treatment of cervical cancer. While small trials have shown comparable survival outcomes between high (HDR) and low-dose rate (LDR) brachytherapy, little data is available in the US. We examined the utilization of HDR brachytherapy and analyzed the impact of type of brachytherapy on survival for cervical cancer. Women with stages IB2-IVA cervical cancer treated with primary (external beam and brachytherapy) radiotherapy between 2003-2011 and recorded in the National Cancer Database (NCDB) were analyzed. Generalized linear mixed models and Cox proportional hazards regression were used to examine predictors of HDR brachytherapy use and the association between HDR use and survival. A total of 10,564 women including 2681 (25.4%) who received LDR and 7883 (74.6%) that received HDR were identified. Use of HDR increased from 50.2% in 2003 to 83.9% in 2011 (Puse of HDR. While patients in the Northeast were more likely to receive HDR therapy, there were no other clinical or socioeconomic characteristics associated with receipt of HDR. In a multivariable Cox model, survival was similar between the HDR and LDR groups (HR=0.93; 95% CI 0.83-1.03). Similar findings were noted in analyses stratified by stage and histology. Kaplan-Meier analyses demonstrated no difference in survival based on type of brachytherapy for stage IIB (P=0.68), IIIB (P=0.17), or IVA (P=0.16) tumors. The use of HDR therapy has increased rapidly. Overall survival is similar for LDR and HDR brachytherapy. Copyright © 2015 Elsevier Inc. All rights reserved.

  6. Low dose rate radiation favors apoptosis as a mechanism of cell death

    International Nuclear Information System (INIS)

    Murtha, Albert D.; Rupnow, Brent; Knox, Susan J.

    1997-01-01

    Purpose/Objective: Radioimmunotherapy (RIT) has demonstrated promising results in the treatment of chemotherapy refractory non-Hodgkin's lymphoma. The radiation associated with this therapy is emitted in a continuous fashion at low dose rates (LDR). Results from studies comparing the relative efficacy of LDR radiation and high dose rate (HDR) radiation on malignant cell killing have been variable. This variability may be due in part to the relative contribution of different mechanisms of cell killing (apoptosis or necrosis) at different dose rates. Materials and Methods: In order to test this hypothesis, the relative efficacy of LDR (16.7 cGy/hr) and HDR radiation (422 cGy/min) were compared using a human B cell lymphoma cell line (PW) and a PW clone (c26) stably transfected to overexpress the anti-apoptotic gene Bcl-2. The endpoints evaluated included the relative amount of cell killing, the fraction of cell killing attributable to apoptosis versus necrosis, and the impact of Bcl-2 overexpression on both overall cell killing and the fraction of killing attributable to apoptosis. Results: HDR and LDR radiation resulted in similar overall cell killing in the PW wild type cell line. In contrast, killing of clone c26 cells was dose rate dependent. One third less killing was seen following LDR irradiation of c26 cells compared with equivalent doses of HDR radiation. Analysis of the relative mechanisms of killing following LDR irradiation revealed a relative increase in the proportion of killing attributable to apoptosis. Conclusion: These findings support the hypothesis that in PW cells, LDR radiation appears to be highly dependent on apoptosis as a mechanism of cell death. These findings may have implications for the selection of patients for RIT, and for the treatment of tumors that overexpress Bcl-2. They may also help form the basis for future rational design of effective combined modality therapies utilizing RIT

  7. Experience of the first application of HDR brachytherapy in nasopharynx

    International Nuclear Information System (INIS)

    Vega Hernandez, Manuel I.; Alfonso Laguardia, Rodolfo; Silvestre Patallo, Ileana; Roca Muchuli, Carlos; Garcia Heredia, Gilda

    2006-01-01

    A research was made by applying boost on the area of the nasopharynx relapse with high dose rate (HDR) in a diagnosis of nasopharynx carcinoma previously treated with telecobalt therapy, at a dose of 70 Gy. There was persistence of the injury. Three sessions were planned, with consecutive fractions of 6.5 Gy in 15 days, with optimization, using a personal mould of autopolymerizable acrylic. The successful possibility to apply the high rate modern brachytherapy was reaffirmed, as a treatment complementary to teletherapy in case of persistence or relapse. A Micro Selectron HDR equipment was used

  8. The use of nomograms in LDR-HDR prostate brachytherapy.

    Science.gov (United States)

    Pujades, Ma Carmen; Camacho, Cristina; Perez-Calatayud, Jose; Richart, José; Gimeno, Jose; Lliso, Françoise; Carmona, Vicente; Ballester, Facundo; Crispín, Vicente; Rodríguez, Silvia; Tormo, Alejandro

    2011-09-01

    The common use of nomograms in Low Dose Rate (LDR) permanent prostate brachytherapy (BT) allows to estimate the number of seeds required for an implant. Independent dosimetry verification is recommended for each clinical dosimetry in BT. Also, nomograms can be useful for dose calculation quality assurance and they could be adapted to High Dose Rate (HDR). This work sets nomograms for LDR and HDR prostate-BT implants, which are applied to three different institutions that use different implant techniques. Patients treated throughout 2010 till April 2011 were considered for this study. This example was chosen to be the representative of the latest implant techniques and to ensure consistency in the planning. A sufficient number of cases for both BT modalities, prescription dose and different work methodology (depending on the institution) were taken into account. The specific nomograms were built using the correlation between the prostate volume and some characteristic parameters of each BT modality, such as the source Air Kerma Strength, number of implanted seeds in LDR or total radiation time in HDR. For each institution and BT modality, nomograms normalized to the prescribed dose were obtained and fitted to a linear function. The parameters of the adjustment show a good agreement between data and the fitting. It should be noted that for each institution these linear function parameters are different, indicating that each centre should construct its own nomograms. Nomograms for LDR and HDR prostate brachytherapy are simple quality assurance tools, specific for each institution. Nevertheless, their use should be complementary to the necessary independent verification.

  9. The use of nomograms in LDR-HDR prostate brachytherapy

    Directory of Open Access Journals (Sweden)

    Ma Carmen Pujades

    2011-09-01

    Full Text Available Purpose: The common use of nomograms in Low Dose Rate (LDR permanent prostate brachytherapy (BT allowsto estimate the number of seeds required for an implant. Independent dosimetry verification is recommended for eachclinical dosimetry in BT. Also, nomograms can be useful for dose calculation quality assurance and they could be adaptedto High Dose Rate (HDR. This work sets nomograms for LDR and HDR prostate-BT implants, which are applied tothree different institutions that use different implant techniques. Material and methods: Patients treated throughout 2010 till April 2011 were considered for this study. This examplewas chosen to be the representative of the latest implant techniques and to ensure consistency in the planning. A sufficientnumber of cases for both BT modalities, prescription dose and different work methodology (depending on theinstitution were taken into account. The specific nomograms were built using the correlation between the prostatevo lume and some characteristic parameters of each BT modality, such as the source Air Kerma Strength, numberof implanted seeds in LDR or total radiation time in HDR. Results: For each institution and BT modality, nomograms normalized to the prescribed dose were obtained andfitted to a linear function. The parameters of the adjustment show a good agreement between data and the fitting.It should be noted that for each institution these linear function parameters are different, indicating that each centreshould construct its own nomograms. Conclusions: Nomograms for LDR and HDR prostate brachytherapy are simple quality assurance tools, specific foreach institution. Nevertheless, their use should be complementary to the necessary independent verification.

  10. Human reliability in high dose rate afterloading radiotherapy based on FMECA

    International Nuclear Information System (INIS)

    Deng Jun; Fan Yaohua; Yue Baorong; Wei Kedao; Ren Fuli

    2012-01-01

    Objective: To put forward reasonable and feasible recommendations against the procedure with relative high risk during the high dose rate (HDR) afterloading radiotherapy, so as to enhance its clinical application safety, through studying the human reliability in the process of carrying out the HDR afterloading radiotherapy. Methods: Basic data were collected by on-site investigation and process analysis as well as expert evaluation. Failure mode, effect and criticality analysis (FMECA) employed to study the human reliability in the execution of HDR afterloading radiotherapy. Results: The FMECA model of human reliability for HDR afterloading radiotherapy was established, through which 25 procedures with relative high risk index were found,accounting for 14.1% of total 177 procedures. Conclusions: FMECA method in human reliability study for HDR afterloading radiotherapy is feasible. The countermeasures are put forward to reduce the human error, so as to provide important basis for enhancing clinical application safety of HDR afterloading radiotherapy. (authors)

  11. High-Dose-Rate Endobronchial Brachytherapy for Recurrent Airway Obstruction From Hyperplastic Granulation Tissue

    International Nuclear Information System (INIS)

    Tendulkar, Rahul D.; Fleming, Peter A.; Reddy, Chandana A.; Gildea, Thomas R.; Machuzak, Michael; Mehta, Atul C.

    2008-01-01

    Purpose: Benign endobronchial granulation tissue causes airway obstruction in up to 20% of patients after lung transplantation or stent placement. High-dose-rate endobronchial brachytherapy (HDR-EB) has been successful in some cases refractory to standard bronchoscopic interventions. Methods and Materials: Between September 2004 and May 2005, 8 patients with refractory benign airway obstruction were treated with HDR-EB, using one to two fractions of Ir-192 prescribed to 7.1 Gy at a radius of 1 cm. Charts were retrospectively reviewed to evaluate subjective clinical response, forced expiratory volume in 1 second (FEV 1 ), and frequency of therapeutic bronchoscopies over 6-month periods before and after HDR-EB. Results: The median follow-up was 14.6 months, and median survival was 10.5 months. The mean number of bronchoscopic interventions improved from 3.1 procedures in the 6-month pretreatment period to 1.8 after HDR-EB. Mean FEV 1 improved from 36% predicted to 46% predicted. Six patients had a good-to-excellent subjective early response, but only one maintained this response beyond 6 months, and this was the only patient treated with HDR-EB within 24 h from the most recent bronchoscopic intervention. Five patients have expired from causes related to their chronic pulmonary disease, including one from hemoptysis resulting from a bronchoarterial fistula. Conclusion: High-dose-rate-EB may be an effective treatment for select patients with refractory hyperplastic granulation tissue causing recurrent airway stenosis. Performing HDR-EB within 24-48 h after excision of obstructive granulation tissue could further improve outcomes. Careful patient selection is important to maximize therapeutic benefit and minimize toxicity. The optimal patient population, dose, and timing of HDR-EB should be investigated prospectively

  12. Comparison of the cost between 60Co and 192Ir, as the sources for high-dose-rate remote control afterloading systems (HDR-RALS)

    International Nuclear Information System (INIS)

    Ogata, Hitoshi

    1994-01-01

    High-Dose-Rate remote control afterloading systems (HDR-RALS) installing 60 Co sources have been prevailing currently in Japan. The survey conducted by Japan Isotope Association (JIA) reports that 180 machines are at working condition. Although the wide prevalence of the HDR-RALS, the stable supply of 60 Co is becoming difficult because of the short availability of raw materials. The supply of 60 Co is planned to be terminated in March 1996. In place of 60 Co, 192 Ir is going to be produced in 1996. The size of 192 Ir, which is much smaller than that of 60 Co, may facilitate broader clinical usability. On the other hand, for the reason that the half life of 192 Ir (73.8 days) is much shorter than that of 60 Co (5.27 years), several exchanges of the sources in a year are necessary. This report analyses the difference of the cost between 60 Co and 192 Ir as the sources for HDR-RALS. As the cost of the 60 Co sources is dependent on the distance from Tokyo. Radiation activity, etc., the cost-calculation was done on the basis the 60 Co sources were installed for the HDR-RALS systems in Yamanashi Central Hospital. The total cost of 60 Co is 3,377,000 yen on the data from JIA. According to the half life of 5.27 years, the available duration can be thought as 7 years and the monthly cost be calculated as about 40,000 yen. In case of 192 Ir, the prices for Buchler' system and Nucletron's system are 800,000 yen and 990,000 yen respectively. Concerning the shortness of the half life, an exchange in every 3 months is ideal. Therefore the monthly cost of 192 Ir would be 260,000-330,000 yen. Consequently the cost-ratio for 192 Ir and 60 Co would become 6.7-8.3. The cost of intracavitary irradiation is controlled by the government as 10,000 yen per treatment in Japan. If this setting remains the same for HDR-RALS installing 192 Ir, almost all the facilities of radiation therapy would suffer from the cost-income inbalance in the near future. (author)

  13. Quality assurance of HDR prostate plans: Program implementation at a community hospital

    International Nuclear Information System (INIS)

    Rush, Jennifer B.; Thomas, Michael D.

    2005-01-01

    Adenocarcinoma of the prostate is currently the most commonly diagnosed cancer in men in the United States, and the second leading cause of cancer mortality. The utilization of radiation therapy is regarded as the definitive local therapy of choice for intermediate- and high-risk disease, in which there is increased risk for extracapsular extension, seminal vesicle invasion, or regional node involvement. High-dose-rate (HDR) brachytherapy is a logical treatment modality to deliver the boost dose to an external beam radiation therapy (EBRT) treatment to increase local control rates. From a treatment perspective, the utilization of a complicated treatment delivery system, the compressed time frame in which the procedure is performed, and the small number of large dose fractions make the implementation of a comprehensive quality assurance (QA) program imperative. One aspect of this program is the QA of the HDR treatment plan. Review of regulatory and medical physics professional publications shows that substantial general guidance is available. We provide some insight to the implementation of an HDR prostate plan program at a community hospital. One aspect addressed is the utilization of the low-dose-rate (LDR) planning system and the use of existing ultrasound image sets to familiarize the radiation therapy team with respect to acceptable HDR implant geometries. Additionally, the use of the LDR treatment planning system provided a means to prospectively determine the relationship between the treated isodose volume and the product of activity and time for the department's planning protocol prior to the first HDR implant. For the first 12 HDR prostate implants, the root-mean-square (RMS) deviation was 3.05% between the predicted product of activity and time vs. the actual plan values. Retrospective re-evaluation of the actual implant data reduced the RMS deviation to 2.36%

  14. Quality assurance of HDR prostate plans: program implementation at a community hospital.

    Science.gov (United States)

    Rush, Jennifer B; Thomas, Michael D

    2005-01-01

    Adenocarcinoma of the prostate is currently the most commonly diagnosed cancer in men in the United States, and the second leading cause of cancer mortality. The utilization of radiation therapy is regarded as the definitive local therapy of choice for intermediate- and high-risk disease, in which there is increased risk for extracapsular extension, seminal vesicle invasion, or regional node involvement. High-dose-rate (HDR) brachytherapy is a logical treatment modality to deliver the boost dose to an external beam radiation therapy (EBRT) treatment to increase local control rates. From a treatment perspective, the utilization of a complicated treatment delivery system, the compressed time frame in which the procedure is performed, and the small number of large dose fractions make the implementation of a comprehensive quality assurance (QA) program imperative. One aspect of this program is the QA of the HDR treatment plan. Review of regulatory and medical physics professional publications shows that substantial general guidance is available. We provide some insight to the implementation of an HDR prostate plan program at a community hospital. One aspect addressed is the utilization of the low-dose-rate (LDR) planning system and the use of existing ultrasound image sets to familiarize the radiation therapy team with respect to acceptable HDR implant geometries. Additionally, the use of the LDR treatment planning system provided a means to prospectively determine the relationship between the treated isodose volume and the product of activity and time for the department's planning protocol prior to the first HDR implant. For the first 12 HDR prostate implants, the root-mean-square (RMS) deviation was 3.05% between the predicted product of activity and time vs. the actual plan values. Retrospective re-evaluation of the actual implant data reduced the RMS deviation to 2.36%.

  15. Vaginal HDR-afterloading in the treatment of endometrial carcinoma - analysis of side effects and relapse rates

    International Nuclear Information System (INIS)

    Weiss, E.; Arnold-Bofinger, H.; Weidner, N.; Hirnle, P.; Bamberg, M.

    1996-01-01

    Introduction: In a retrospective analysis side effects, relapse rates and the value of in vivo dosimetry for vaginal HDR-brachytherapy (BT) in women with endometrial carcinoma are evaluated. Material and Methods: From 1987 to 1993 124 women received vaginal HDR-BT (Ir-192) only for adjuvant treatment of early endometrial carcinoma. The usual fractionation was 1x7 Gy per week applied to the surface of the vaginal cylinder to a total dose of 21 Gy. During BT we performed in vivo dosimetry using bladder and rectal probes. The measured doses per patient in the bladder ranged from 1.5 to 9.8 Gy, 6.7 Gy in the mean. In the rectum the mean dose was 8.2 Gy (1.7 to 13.8 Gy). Results: The analysis of side effects shows that only 23 of 124 women (18 %) reported any problems at all. 13 women (10 %) noticed dysuria or pollakisuria for a maximum of 6 weeks. The average measured bladder dose in those patients was 5.3 Gy, (below the average bladder dose of 6.7 Gy). Diarrhea was reported in 3 patients (2 %), in those the average measured rectal dose was 6.8 Gy (lower than the average rectal dose of 8.2 Gy). 6 women (5 %) received antibiotic treatment because of bacteriuria. Of all patients 10 (8%) had a pelvic relapse, 3 in the vagina only. 5 women showed distant metastases during follow-up (3 of those having also a local relapse). Discussion and Conclusion: With only 2 % vaginal relapses HDR-BT is a save method to reduce the probability of vaginal metastases of endometrial carcinoma in an adjuvant setting. Moreover the incidence of acute side effects is low, no lasting complications were seen in our patients. There existed no correlation between the incidence of side effects and the measured in vivo doses in bladder and rectum, questioning the future usage in the era of well advanced BT-planning systems

  16. MO-B-BRC-00: Prostate HDR Treatment Planning - Considering Different Imaging Modalities

    International Nuclear Information System (INIS)

    2016-01-01

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions

  17. MO-B-BRC-00: Prostate HDR Treatment Planning - Considering Different Imaging Modalities

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2016-06-15

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions.

  18. High-dose-rate intracavitary brachytherapy in the management of cervical and vaginal intraepithelial neoplasia

    International Nuclear Information System (INIS)

    Ogino, Ichiro; Kitamura, Tatsuo; Okajima, Hiroyuki; Matsubara, Sho

    1998-01-01

    Purpose: To assess the effectiveness of high-dose rate intracavitary brachytherapy (HDR-ICR) in patients with grade 3 cervical intraepithelial neoplasia (CIN-3) and grade 3 vaginal intraepithelial neoplasia (VAIN-3). Methods and Materials: This was a retrospective analysis in 20 patients with CIN-3 (n = 14) or VAIN-3 (n = 6), average age 61.9 years, managed with HDR-ICR at Kanagawa Cancer Center. Two patients with CIN-3 with microinvasive foci and 11 other patients with CIN-3 were treated with HDR-ICR for cervical lesions. Six patients with CIN-3 after hysterectomy received HDR-ICR for recurrent or residual VAIN-3 lesions. One patient received radiation therapy for both CIN-3 and VAIN-3 lesions. All these patients but one were postmenopausal. Results: Seventeen patients were treated with HDR-ICR alone, and three with combined external radiation therapy. The dose was calculated at Point A located 2 cm superior to the external os and 2 cm lateral to the axis of the intrauterine tube for intact uterus. For lesions of the vaginal stump, the dose was calculated at a point 1 cm superior to the vaginal apex or 1 cm beyond vaginal mucosa. In the 14 patients treated for CIN-3 lesions, the mean total dose of HDR-ICR was 26.1 Gy (range 20-30). Six patients received HDR-ICR for VAIN-3 lesions with mean dose of 23.3 Gy (range 15-30). At follow-up (mean 90.5 months; range 13-153), 14 patients were alive and 6 had died owing to nonmalignant intercurrent disease. No patient developed recurrent disease. Rectal bleeding occurred in three patients, but this symptom subsided spontaneously. Moderate and severe vaginal reactions were noted in two patients, in whom the treatment had included the entire vagina. Conclusions: HDR-ICR can be employed as the primary management strategy for postmenopausal women with CIN-3. In intraepithelial neoplasia involving the vaginal wall after hysterectomy, HDR-ICR should be considered as an alternative to total vaginectomy

  19. The relative efficacy of HDR and LDR interstitial brachytherapy in squamous cell carcinoma of vagina

    International Nuclear Information System (INIS)

    Demanes, D. J.; Hsu, I-C.; Lin, S.; Ewing, T.; Rodriguez, R.

    1996-01-01

    Introduction: Beginning in 1982 we performed low dose rate (LDR) interstitial template brachytherapy (ISTB) for carcinoma of the vagina. High dose rate (HDR) remote afterloading has been used exclusively since 1991. We compare the results LDR and HDR brachytherapy. Material and Methods: Between 1982 and 1994, 30 patients with primary squamous cell carcinoma of vagina received external beam radiotherapy (EBRT) and brachytherapy. The AJCC stage distribution was 3 stage I, 25 stage II, and 2 stage III. The average central pelvic EBRT dose was 35 Gy. Pelvic side wall EBRT doses ranged from 45 to 50.4 Gy. Nineteen patients had LDR treatment; 3 intracavitary brachytherapy (ICB) and 16 ISTB. Eleven patients had HDR treatment; 2 ICB and 9 ISB. The average dose delivered by LDR was 41.2 Gy usually in 2 fractions, and by HDR 32.5 Gy in 6 fractions of 500-550 cGy. Local failures were confirmed pathologically. The absolute survival (AS) and relapse-free survival (RFS) were calculated using Kaplan-Meier method and compared with logrank statistics. Results: The mean follow-up was 77 months for LDR and 23 months for HDR. Local and regional control was achieved in 90% (27/30) of the patients. Three year AS was 84% and RFS was 87%. There was no significant difference between LDR and HDR in AS, RFS or local-regional control, (log rank p=0.85, p=0.12 and p=0.35 respectively). The single HDR local failure presented in a patient with extensive stage II disease who declined ISTB. There were fewer complications following HDR. The 1 case of extensive vaginal necrosis and the 3 cases of rectovaginal fistula that required surgery occurred only with LDR brachytherapy. Discussion: Excellent local and regional control of carcinoma of the vagina can be achieved by administering limited doses of external radiation and brachytherapy. Interstitial template implants are the best means of encompassing paravaginal disease while sparing the adjacent uninvolved normal tissues from high doses of

  20. CONTRASTING DOSE-RATE EFFECTS OF GAMMA-IRRADIATION ON RAT SALIVARY-GLAND FUNCTION

    NARCIS (Netherlands)

    VISSINK, A; DOWN, JD; KONINGS, AWT

    The aim of this study was to investigate the effects of Co-60 irradiation delivered at high (HDR) and low (LDR) dose-rates on rat salivary gland function. Total-body irradiation (TBI; total doses 7.5, 10 and 12.5 Gy) was applied from a Co-60 source at dose-rates of 1 cGy/min (LDR) and 40 cGy/min

  1. Dosimetric Considerations to Determine the Optimal Technique for Localized Prostate Cancer Among External Photon, Proton, or Carbon-Ion Therapy and High-Dose-Rate or Low-Dose-Rate Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Georg, Dietmar, E-mail: Dietmar.Georg@akhwien.at [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Hopfgartner, Johannes [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Gòra, Joanna [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Kuess, Peter [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Kragl, Gabriele [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Berger, Daniel [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Hegazy, Neamat [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Goldner, Gregor; Georg, Petra [Department of Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria); Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna/Allgemeines Krankenhaus der Stadt Wien, Vienna (Austria)

    2014-03-01

    Purpose: To assess the dosimetric differences among volumetric modulated arc therapy (VMAT), scanned proton therapy (intensity-modulated proton therapy, IMPT), scanned carbon-ion therapy (intensity-modulated carbon-ion therapy, IMIT), and low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy (BT) treatment of localized prostate cancer. Methods and Materials: Ten patients were considered for this planning study. For external beam radiation therapy (EBRT), planning target volume was created by adding a margin of 5 mm (lateral/anterior–posterior) and 8 mm (superior–inferior) to the clinical target volume. Bladder wall (BW), rectal wall (RW), femoral heads, urethra, and pelvic tissue were considered as organs at risk. For VMAT and IMPT, 78 Gy(relative biological effectiveness, RBE)/2 Gy were prescribed. The IMIT was based on 66 Gy(RBE)/20 fractions. The clinical target volume planning aims for HDR-BT ({sup 192}Ir) and LDR-BT ({sup 125}I) were D{sub 90%} ≥34 Gy in 8.5 Gy per fraction and D{sub 90%} ≥145 Gy. Both physical and RBE-weighted dose distributions for protons and carbon-ions were converted to dose distributions based on 2-Gy(IsoE) fractions. From these dose distributions various dose and dose–volume parameters were extracted. Results: Rectal wall exposure 30-70 Gy(IsoE) was reduced for IMIT, LDR-BT, and HDR-BT when compared with VMAT and IMPT. The high-dose region of the BW dose–volume histogram above 50 Gy(IsoE) of IMPT resembled the VMAT shape, whereas all other techniques showed a significantly lower high-dose region. For all 3 EBRT techniques similar urethra D{sub mean} around 74 Gy(IsoE) were obtained. The LDR-BT results were approximately 30 Gy(IsoE) higher, HDR-BT 10 Gy(IsoE) lower. Normal tissue and femoral head sparing was best with BT. Conclusion: Despite the different EBRT prescription and fractionation schemes, the high-dose regions of BW and RW expressed in Gy(IsoE) were on the same order of magnitude. Brachytherapy techniques

  2. Dosimetric Considerations to Determine the Optimal Technique for Localized Prostate Cancer Among External Photon, Proton, or Carbon-Ion Therapy and High-Dose-Rate or Low-Dose-Rate Brachytherapy

    International Nuclear Information System (INIS)

    Georg, Dietmar; Hopfgartner, Johannes; Gòra, Joanna; Kuess, Peter; Kragl, Gabriele; Berger, Daniel; Hegazy, Neamat; Goldner, Gregor; Georg, Petra

    2014-01-01

    Purpose: To assess the dosimetric differences among volumetric modulated arc therapy (VMAT), scanned proton therapy (intensity-modulated proton therapy, IMPT), scanned carbon-ion therapy (intensity-modulated carbon-ion therapy, IMIT), and low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy (BT) treatment of localized prostate cancer. Methods and Materials: Ten patients were considered for this planning study. For external beam radiation therapy (EBRT), planning target volume was created by adding a margin of 5 mm (lateral/anterior–posterior) and 8 mm (superior–inferior) to the clinical target volume. Bladder wall (BW), rectal wall (RW), femoral heads, urethra, and pelvic tissue were considered as organs at risk. For VMAT and IMPT, 78 Gy(relative biological effectiveness, RBE)/2 Gy were prescribed. The IMIT was based on 66 Gy(RBE)/20 fractions. The clinical target volume planning aims for HDR-BT ( 192 Ir) and LDR-BT ( 125 I) were D 90% ≥34 Gy in 8.5 Gy per fraction and D 90% ≥145 Gy. Both physical and RBE-weighted dose distributions for protons and carbon-ions were converted to dose distributions based on 2-Gy(IsoE) fractions. From these dose distributions various dose and dose–volume parameters were extracted. Results: Rectal wall exposure 30-70 Gy(IsoE) was reduced for IMIT, LDR-BT, and HDR-BT when compared with VMAT and IMPT. The high-dose region of the BW dose–volume histogram above 50 Gy(IsoE) of IMPT resembled the VMAT shape, whereas all other techniques showed a significantly lower high-dose region. For all 3 EBRT techniques similar urethra D mean around 74 Gy(IsoE) were obtained. The LDR-BT results were approximately 30 Gy(IsoE) higher, HDR-BT 10 Gy(IsoE) lower. Normal tissue and femoral head sparing was best with BT. Conclusion: Despite the different EBRT prescription and fractionation schemes, the high-dose regions of BW and RW expressed in Gy(IsoE) were on the same order of magnitude. Brachytherapy techniques were clearly superior in

  3. The effect of low dose rate on metabolomic response to radiation in mice

    International Nuclear Information System (INIS)

    Goudarzi, Maryam; Mak, Tytus D.; Chen, Congju; Smilenov, Lubomir B.; Brenner, David J.; Fornace, Albert J.

    2014-01-01

    Metabolomics has been shown to have utility in assessing responses to exposure by ionizing radiation (IR) in easily accessible biofluids such as urine. Most studies to date from our laboratory and others have employed γ-irradiation at relatively high dose rates (HDR), but many environmental exposure scenarios will probably be at relatively low dose rates (LDR). There are well-documented differences in the biologic responses to LDR compared to HDR, so an important question is to assess LDR effects at the metabolomics level. Our study took advantage of a modern mass spectrometry approach in exploring the effects of dose rate on the urinary excretion levels of metabolites 2 days after IR in mice. A wide variety of statistical tools were employed to further focus on metabolites, which showed responses to LDR IR exposure (0.00309 Gy/min) distinguishable from those of HDR. From a total of 709 detected spectral features, more than 100 were determined to be statistically significant when comparing urine from mice irradiated with 1.1 or 4.45 Gy to that of sham-irradiated mice 2 days post-exposure. The results of this study show that LDR and HDR exposures perturb many of the same pathways such as TCA cycle and fatty acid metabolism, which also have been implicated in our previous IR studies. However, it is important to note that dose rate did affect the levels of particular metabolites. Differences in urinary excretion levels of such metabolites could potentially be used to assess an individual's exposure in a radiobiological event and thus would have utility for both triage and injury assessment. (orig.)

  4. High-dose-rate brachytherapy in uterine cervical carcinoma

    International Nuclear Information System (INIS)

    Patel, Firuza D.; Rai, Bhavana; Mallick, Indranil; Sharma, Suresh C.

    2005-01-01

    Purpose: High-dose-rate (HDR) brachytherapy is in wide use for curative treatment of cervical cancer. The American Brachytherapy Society has recommended that the individual fraction size be <7.5 Gy and the range of fractions should be four to eight; however, many fractionation schedules, varying from institution to institution, are in use. We use 9 Gy/fraction of HDR in two to five fractions in patients with carcinoma of the uterine cervix. We found that our results and toxicity were comparable to those reported in the literature and hereby present our experience with this fractionation schedule. Methods and Materials: A total of 121 patients with Stage I-III carcinoma of the uterine cervix were treated with HDR brachytherapy between 1996 and 2000. The total number of patients analyzed was 113. The median patient age was 53 years, and the histopathologic type was squamous cell carcinoma in 93% of patients. The patients were subdivided into Groups 1 and 2. In Group 1, 18 patients with Stage Ib-IIb disease, tumor size <4 cm, and preserved cervical anatomy underwent simultaneous external beam radiotherapy to the pelvis to a dose of 40 Gy in 20 fractions within 4 weeks with central shielding and HDR brachytherapy of 9 Gy/fraction, given weekly, and interdigitated with external beam radiotherapy. The 95 patients in Group 2, who had Stage IIb-IIIb disease underwent external beam radiotherapy to the pelvis to a dose of 46 Gy in 23 fractions within 4.5 weeks followed by two sessions of HDR intracavitary brachytherapy of 9 Gy each given 1 week apart. The follow-up range was 3-7 years (median, 36.4 months). Late toxicity was graded according to the Radiation Therapy Oncology Group criteria. Results: The 5-year actuarial local control and disease-free survival rate was 74.5% and 62.0%, respectively. The actuarial local control rate at 5 years was 100% for Stage I, 80% for Stage II, and 67.2% for Stage III patients. The 5-year actuarial disease-free survival rate was 88.8% for

  5. SU-F-BRA-04: Prostate HDR Brachytherapy with Multichannel Robotic System

    International Nuclear Information System (INIS)

    Joseph, F Maria; Podder, T; Yu, Y

    2015-01-01

    Purpose: High-dose-rate (HDR) brachytherapy is gradually becoming popular in treating patients with prostate cancers. However, placement of the HDR needles at desired locations into the patient is challenging. Application of robotic system may improve the accuracy of the clinical procedure. This experimental study is to evaluate the feasibility of using a multichannel robotic system for prostate HDR brachytherapy. Methods: In this experimental study, the robotic system employed was a 6-DOF Multichannel Image-guided Robotic Assistant for Brachytherapy (MIRAB), which was designed and fabricated for prostate seed implantation. The MIRAB has the provision of rotating 16 needles while inserting them. Ten prostate HDR brachytherapy needles were simultaneously inserted using MIRAB into a commercially available prostate phantom. After inserting the needles into the prostate phantom at desired locations, 2mm thick CT slices were obtained for dosimetric planning. HDR plan was generated using Oncetra planning system with a total prescription dose of 34Gy in 4 fractions. Plan quality was evaluated considering dose coverage to prostate and planning target volume (PTV), with 3mm margin around prostate, as well as the dose limit to the organs at risk (OARs) following the American Brachytherapy Society (ABS) guidelines. Results: From the CT scan, it is observed that the needles were inserted straight into the desired locations and they were adequately spaced and distributed for a clinically acceptable HDR plan. Coverage to PTV and prostate were about 91% (V100= 91%) and 96% (V100=96%), respectively. Dose to 1cc of urethra, rectum, and bladder were within the ABS specified limits. Conclusion: The MIRAB was able to insert multiple needles simultaneously into the prostate precisely. By controlling the MIRAB to insert all the ten utilized needles into the prostate phantom, we could achieve the robotic HDR brachytherapy successfully. Further study for assessing the systemâ

  6. Transition from LDR to HDR brachytherapy for cervical cancer: Evaluation of tumor control, survival, and toxicity.

    Science.gov (United States)

    Romano, K D; Pugh, K J; Trifiletti, D M; Libby, B; Showalter, T N

    In 2012, our institution transitioned from low-dose-rate (LDR) brachytherapy to high dose-rate (HDR) brachytherapy. We report clinical outcomes after brachytherapy for cervical cancer at our institution over a continuous 10-year period. From 2004 to 2014, 258 women (184 LDR and 74 HDR) were treated with tandem and ovoid brachytherapy in the multidisciplinary management of International Federation of Gynecology and Obstetrics Stages IA-IVB cervical cancer. Clinical and treatment-related prognostic factors including age, stage, smoking status, relevant doses, and toxicity data were recorded. Median followup for the LDR and HDR groups was 46 months and 12 months, respectively. The majority of patients (92%) received external beam radiotherapy as well as concurrent chemotherapy (83%) before the start of brachytherapy. For all stages, the 1-year local control and overall survival (OS) rates were comparable between the LDR and HDR groups (87% vs. 81%, p = 0.12; and 75% vs. 85%, p = 0.16), respectively. Factors associated with OS on multivariate analysis include age, stage, and nodal involvement. On multivariate analysis, severe toxicity (acute or chronic) was higher with HDR than LDR (24% vs. 10%, p = 0.04). Additional prognostic factors associated with increased severe toxicity include former/current smokers and total dose to lymph nodes. This comparative retrospective analysis of a large cohort of women treated with brachytherapy demonstrates no significant difference in OS or local control between the LDR and HDR. Acute and chronic toxicity increased shortly after the implementation of HDR, highlighting the importance of continued refinement of HDR methods, including integrating advanced imaging. Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  7. Comparison of low and high dose rate brachytherapy in the treatment of uterine cervix cancer. Retrospective analysis of two sequential series

    International Nuclear Information System (INIS)

    Ferrigno, Robson; Nishimoto, Ines Nobuko; Ribeiro dos Santos Novaes, Paulo Eduardo; Pellizzon, Antonio Cassio Assis; Conte Maia, Maria Aparecida; Fogarolli, Ricardo Cesar; Salvajoli, Joao Victor

    2005-01-01

    Purpose: This retrospective analysis aims to report on the comparative outcome of cervical cancer patients treated with low dose rate (LDR) and high dose rate (HDR) brachytherapy. Methods and Materials: From 1989 to 1995, 190 patients were treated with low dose rate (LDR) brachytherapy (LDR group) and from 1994 to 2001, 118 patients were treated with high dose rate (HDR) brachytherapy (HDR group). FIGO stage distribution for the LDR group was Stage I: 6.3%; Stage II: 57.4%; and Stage III: 36.3% and for the HDR group Stage I: 9.3%; Stage II: 43.2%; and Stage III: 47.4%. All patients were treated with telecobalt external-beam radiotherapy (EBR). Median doses of LDR brachytherapy at Point A were 40 Gy and 50 Gy for patients treated with 1 and 2 implants, respectively. All patients from the HDR group were treated with 24 Gy in 4 fractions of 6 Gy to Point A. Survival, disease-free survival, local control, and late complications at 5 years, were endpoints compared for both groups. Results: Median follow-up time for LDR and HDR groups was 70 months (range, 8-127 months) and 33 months (range, 4-117 months), respectively. For all stages combined, overall survival, disease-free survival, and local control at 5 years were better in the LDR group (69% vs. 55%, p = 0.007; 73% vs. 56%, p = 0.002; and 74% vs. 65%; p = 0.04, respectively). For clinical Stages I and II, no differences was seen in overall survival, disease-free survival, and local control at 5 years between the two groups. For clinical Stage III, overall survival and disease-free survival at 5 years were better in the LDR group than in the HDR group (46% vs. 36%, p = 0.04 and 49% vs. 37%, p = 0.03, respectively), and local control was marginally higher in the LDR group than in the HDR group (58% vs. 50%, p = 0.19). The 5-year probability of rectal complications was higher in the LDR group than in the HDR group (16% vs. 8%, p = 0.03) and 5-year probability of small bowel and urinary complications was not

  8. Evaluation of Wall Correction Factor of INER's Air-Kerma Primary Standard Chamber and Dose Variation by Source Displacement for HDR 192Ir Brachytherapy

    Directory of Open Access Journals (Sweden)

    J. H. Lee

    2013-01-01

    Full Text Available The aim of the present study was to estimate the wall effect of the self-made spherical graphite-walled cavity chamber with the Monte Carlo method for establishing the air-kerma primary standard of high-dose-rate (HDR 192Ir brachytherapy sources at the Institute of Nuclear Energy Research (INER, Taiwan. The Monte Carlo method established in this paper was also employed to respectively simulate wall correction factors of the 192Ir air-kerma standard chambers used at the National Institute of Standards and Technology (NIST, USA and the National Physical Laboratory (NPL, UK for comparisons and verification. The chamber wall correction calculation results will be incorporated into INER's HDR 192Ir primary standard in the future. For the brachytherapy treatment in the esophagus or in the bronchi, the position of the isotope may have displacement in the cavity. Thus the delivered dose would differ from the prescribed dose in the treatment plan. We also tried assessing dose distribution due to the position displacement of HDR 192Ir brachytherapy source in a phantom with a central cavity by the Monte Carlo method. The calculated results could offer a clinical reference for the brachytherapy within the human organs with cavity.

  9. Phase III trial of high and low dose rate interstitial radiotherapy for early oral tongue cancer

    International Nuclear Information System (INIS)

    Inoue, Takehiro; Inoue, Toshihiko; Teshima, Teruki; Murayama, Shigeyuki; Shimizutani, Kimishige; Fuchihata, Hajime; Furukawa, Souhei

    1996-01-01

    Purpose: Oral tongue carcinomas are highly curable with radiotherapy. In the past, patients with tongue carcinoma have usually been treated with low dose rate (LDR) interstitial radiation. This Phase III study was designed to compare the treatment results obtained with LDR with those obtained with high dose rate (HDR) interstitial radiotherapy for tongue carcinoma. Methods and Materials: The criteria for patient selection for the Phase III study were: (a) presence of a T1T2N0 tumor that could be treated with single-plane implantation, (b) localization of tumor at the lateral tongue border, (c) tumor thickness of 10 mm or less, (d) performance status between O and 3, and (e) absence of any severe concurrent disease. From April 1992 through December 1993, 15 patients in the LDR group (70 Gy/4 to 9 days) and 14 patients in the HDR group (60 Gy/10 fractions/6 days) were accrued. The time interval between two fractions of the HDR brachytherapy was more than 6 h. Results: Local recurrence occurred in two patients treated with LDR brachytherapy but in none of the patients treated with HDR. One- and 2-year local control rates for patients in the LDR group were both 86%, compared with 100% in the HDR group (p = 0.157). There were four patients with nodal metastasis in the LDR group and three in the HDR group. Local recurrence occurred in two of the four patients with nodal metastases in the LDR group. One- and 2-year nodal control rates for patients in the LDR group are were 85%, compared with 79% in the HDR group. Conclusion: HDR fractionated interstitial brachytherapy can be an alternative to traditional LDR brachytherapy for early tongue cancer and eliminate the radiation exposure for medical staffs

  10. Impact of catheter reconstruction error on dose distribution in high dose rate intracavitary brachytherapy and evaluation of OAR doses

    International Nuclear Information System (INIS)

    Thaper, Deepak; Shukla, Arvind; Rathore, Narendra; Oinam, Arun S.

    2016-01-01

    In high dose rate brachytherapy (HDR-B), current catheter reconstruction protocols are relatively slow and error prone. The purpose of this study is to evaluate the impact of catheter reconstruction error on dose distribution in CT based intracavitary brachytherapy planning and evaluation of its effect on organ at risk (OAR) like bladder, rectum and sigmoid and target volume High risk clinical target volume (HR-CTV)

  11. High dose rate brachytherapy for superficial cancer of the esophagus

    International Nuclear Information System (INIS)

    Maingon, Philippe; D'Hombres, Anne; Truc, Gilles; Barillot, Isabelle; Michiels, Christophe; Bedenne, Laurent; Horiot, Jean Claude

    2000-01-01

    Purpose: We analyzed our experience with external radiotherapy, combined modality treatment, or HDR brachytherapy alone to limited esophageal cancers. Methods and Materials: From 1991 to 1996, 25 patients with limited superficial esophagus carcinomas were treated by high dose rate brachytherapy. The mean age was 63 years (43-86 years). Five patients showed superficial local recurrence after external radiotherapy. Eleven patients without invasion of the basal membrane were staged as Tis. Fourteen patients with tumors involving the submucosa without spreading to the muscle were staged as T1. Treatment consisted of HDR brachytherapy alone in 13 patients, external radiotherapy and brachytherapy in 8 cases, and concomitant chemo- and radiotherapy in 4 cases. External beam radiation was administered to a total dose of 50 Gy using 2 Gy daily fractions in 5 weeks. In cases of HDR brachytherapy alone (13 patients), 6 applications were performed once a week. Results: The mean follow-up is 31 months (range 24-96 months). Twelve patients received 2 applications and 13 patients received 6 applications. Twelve patients experienced a failure (48%), 11/12 located in the esophagus, all of them in the treated volume. One patient presented an isolated distant metastasis. In the patients treated for superficial recurrence, 4/5 were locally controlled (80%) by brachytherapy alone. After brachytherapy alone, 8/13 patients were controlled (61%). The mean disease-free survival is 14 months (1-36 months). Overall survival is 76% at 1 year, 37% at 2 years, and 14% at 3 years. Overall survival for Tis patients is 24% vs. 20% for T1 (p 0.83). Overall survival for patients treated by HDR brachytherapy alone is 43%. One patient presented with a fistula with local failure after external radiotherapy and brachytherapy. Four stenosis were registered, two were diagnosed on barium swallowing without symptoms, and two required dilatations. Conclusion: High dose rate brachytherapy permits the treating

  12. HDRMC, an accelerated Monte Carlo dose calculator for high dose rate brachytherapy with CT-compatible applicators

    Energy Technology Data Exchange (ETDEWEB)

    Chibani, Omar, E-mail: omar.chibani@fccc.edu; C-M Ma, Charlie [Fox Chase Cancer Center, Philadelphia, Pennsylvania 19111 (United States)

    2014-05-15

    Purpose: To present a new accelerated Monte Carlo code for CT-based dose calculations in high dose rate (HDR) brachytherapy. The new code (HDRMC) accounts for both tissue and nontissue heterogeneities (applicator and contrast medium). Methods: HDRMC uses a fast ray-tracing technique and detailed physics algorithms to transport photons through a 3D mesh of voxels representing the patient anatomy with applicator and contrast medium included. A precalculated phase space file for the{sup 192}Ir source is used as source term. HDRM is calibrated to calculated absolute dose for real plans. A postprocessing technique is used to include the exact density and composition of nontissue heterogeneities in the 3D phantom. Dwell positions and angular orientations of the source are reconstructed using data from the treatment planning system (TPS). Structure contours are also imported from the TPS to recalculate dose-volume histograms. Results: HDRMC was first benchmarked against the MCNP5 code for a single source in homogenous water and for a loaded gynecologic applicator in water. The accuracy of the voxel-based applicator model used in HDRMC was also verified by comparing 3D dose distributions and dose-volume parameters obtained using 1-mm{sup 3} versus 2-mm{sup 3} phantom resolutions. HDRMC can calculate the 3D dose distribution for a typical HDR cervix case with 2-mm resolution in 5 min on a single CPU. Examples of heterogeneity effects for two clinical cases (cervix and esophagus) were demonstrated using HDRMC. The neglect of tissue heterogeneity for the esophageal case leads to the overestimate of CTV D90, CTV D100, and spinal cord maximum dose by 3.2%, 3.9%, and 3.6%, respectively. Conclusions: A fast Monte Carlo code for CT-based dose calculations which does not require a prebuilt applicator model is developed for those HDR brachytherapy treatments that use CT-compatible applicators. Tissue and nontissue heterogeneities should be taken into account in modern HDR

  13. Effect of radiation dose rate and cyclophosphamide on pulmonary toxicity after total body irradiation in a mouse model

    International Nuclear Information System (INIS)

    Safwat, Akmal; Nielsen, Ole S.; El-Badawy, Samy; Overgaard, Jens

    1996-01-01

    Purpose: Interstitial pneumonitis (IP) is still a major complication after total body irradiation (TBI) and bone marrow transplantation (BMT). It is difficult to determine the exact role of radiation in this multifactorial complication, especially because most of the experimental work on lung damage was done using localized lung irradiation and not TBI. We have thus tested the effect of radiation dose rate and combining cyclophosphamide (CTX) with single fraction TBI on lung damage in a mouse model for BMT. Methods and Materials: TBI was given as a single fraction at a high dose rate (HDR, 0.71 Gy/min) or a low dose rate (LDR, 0.08 Gy/min). CTX (250 mg/kg) was given 24 h before TBI. Bone marrow transplantation (BMT) was performed 4-6 h after the last treatment. Lung damage was assessed using ventilation rate (VR) and lethality between 28 and 180 days (LD (50(28))-180 ). Results: The LD 50 for lung damage, ± standard error (SE), increased from 12.0 (± 0.2) Gy using single fraction HDR to 15.8 (± 0.6) Gy using LDR. Adding CTX shifted the dose-response curves towards lower doses. The LD 50 values for the combined treatment were 5.3 (± 0.2) and 3.5 (± 0.2) Gy for HDR and LDR, respectively. This indicates that the combined effect of CTX and LDR was more toxic than that of combined CTX and HDR. Lung damage evaluated by VR demonstrated two waves of VR increase. The first wave of VR increase occurred after 6 weeks using TBI only and after 3 weeks in the combined CTX-TBI treatment, irrespective of total dose or dose rate. The second wave of VR elevation resembled the IP that follows localized thoracic irradiation in its time of occurrence. Conclusions: Lung damage following TBI could be spared using LDR. However, CTX markedly enhances TBI-induced lung damage. The combination of CTX and LDR is more toxic to the lungs than combining CTX and HDR

  14. Dose volume assessment of high dose rate 192IR endobronchial implants

    International Nuclear Information System (INIS)

    Cheng, B. Saw; Korb, Leroy J.; Pawlicki, Todd; Wu, Andrew

    1996-01-01

    Purpose: To study the dose distributions of high dose rate (HDR) endobronchial implants using the dose nonuniformity ratio (DNR) and three volumetric irradiation indices. Methods and Materials: Multiple implants were configured by allowing a single HDR 192 Ir source to step through a length of 6 cm along an endobronchial catheter. Dwell times were computed to deliver a dose of 5 Gy to points 1 cm away from the catheter axis. Five sets of source configurations, each with different dwell position spacings from 0.5 to 3.0 cm, were evaluated. Three-dimensional (3D) dose distributions were then generated for each source configuration. Differential and cumulative dose-volume curves were generated to quantify the degree of target volume coverage, dose nonuniformity within the target volume, and irradiation of tissues outside the target volume. Evaluation of the implants were made using the DNR and three volumetric irradiation indices. Results: The observed isodose distributions were not able to satisfy all the dose constraints. The ability to optimally satisfy the dose constraints depended on the choice of dwell position spacing and the specification of the dose constraint points. The DNR and irradiation indices suggest that small dwell position spacing does not result in a more homogeneous dose distribution for the implant. This study supports the existence of a relationship between the dwell position spacing and the distance from the catheter axis to the reference dose or dose constraint points. Better dose homogeneity for an implant can be obtained if the spacing of the dwell positions are about twice the distance from the catheter axis to the reference dose or dose constraint points

  15. Simultaneous radiochemotherapy and endoluminal HDR brachytherapy in esophageal cancer; Simultane Radiochemotherapie mit intraluminaler HDR-Brachytherapie des Oesophaguskarzinoms

    Energy Technology Data Exchange (ETDEWEB)

    Patonay, P.; Naszaly, A.; Mayer, A. [Hauptstaedtisches Zentrum fuer Radioonkologie und Strahlentherapie, Budapest (Hungary)

    2007-02-15

    Purpose: to study efficacy and toxicity of radiochemotherapy in esophageal cancer including initial endoluminal high-dose-rate brachytherapy (HDR-BT). Patients and methods: between 01/1995 and 06/2005, 61 patients with esophageal cancer were treated preoperatively with definitive and palliative intent. Treatment started with intraluminal HDR-BT for recanalization of the esophagus (single fraction size of 8 Gy in 0.5 cm depth, three times, q7d) followed by external-beam radiation therapy (50 Gy total dose, 5 x 2 Gy/week, 25 fractions in 5 weeks). Chemotherapy was started simultaneously with external irradiation (three courses of cisplatin and 5-fluorouracil, q21d). Results: swallowing function improved in 55/61 patients (dysphagia classification according to the RTOG), and worsened in 6/61 patients, respectively. Median duration of symptomatic improvement was 11 months, median follow-up 12 months (range 3-68 months). Following simultaneous radiochemotherapy, tumor resectability was achieved in 7/25 patients of the neoadjuvant group, and the histological specimen showed complete remission in 6/7 patients. Conclusion: these results indicate a favorable effect of simultaneous radiochemotherapy starting with endoluminal HDR-after-loading-(AL-)BT in esophageal cancer. (orig.)

  16. Accurate assessment of the distortions produced by the transit dose in HDR brachytherapy

    International Nuclear Information System (INIS)

    Nani, E.K.; Kyere, A.W.K.; Tetteh, K.

    2001-01-01

    Current polynomial methods used in the modelling of the dose distributions in HDR brachytherapy have been reformulated to improve accuracy. An example is provided to show the effects of the transit dose on the output. The transit dose, which is neglected by current computer software for calculating doses, can result in significant dosimetric errors. These additional unrecognised doses imply over-dosing and distortions in the dose distributions within the irradiated volume. Assessment of dose to critical and radiosensitive organs is therefore inaccurate. These could increase late tissue complications as predicted by the Linear Quadratic Model. Our model works very well for straight catheters and is highly recommended for the evaluation of the transit dose around such catheters. (author)

  17. Assessing patient characteristics and radiation-induced non-targeted effects in vivo for high dose-rate (HDR) brachytherapy.

    Science.gov (United States)

    Pinho, Christine; Timotin, Emilia; Wong, Raimond; Sur, Ranjan K; Hayward, Joseph E; Farrell, Thomas J; Seymour, Colin; Mothersill, Carmel

    2015-01-01

    To test whether blood, urine, and tissue based colony-forming assays are a useful clinical detection tool for assessing fractionated treatment responses and non-targeted radiation effects in bystander cells. To assess patients' responses to radiation treatments, blood serum, urine, and an esophagus explant-based in vivo colony-forming assay were used from oesophageal carcinoma patients. These patients underwent three fractions of high dose rate (HDR) intraluminal brachytherapy (ILBT). Human keratinocyte reporters exposed to blood sera taken after the third fraction of brachytherapy had a significant increase in cloning efficiency compared to baseline samples (p fractions for the blood sera data only. Patient characteristics such as gender had no statistically significant effect (p > 0.05). Large variability was observed among the patients' tissue samples, these colony-forming assays showed no significant changes throughout fractionated brachytherapy (p > 0.05). Large inter-patient variability was found in the urine and tissue based assays, so these techniques were discontinued. However, the simple blood-based assay had much less variability. This technique may have future applications as a biological dosimeter to predict treatment outcome and assess non-targeted radiation effects.

  18. High-dose-rate intraoperative radiation therapy: the nuts and bolts of starting a program

    Science.gov (United States)

    Moningi, Shalini; Armour, Elwood P.; Terezakis, Stephanie A.; Efron, Jonathan E.; Gearhart, Susan L.; Bivalacqua, Trinity J.; Kumar, Rachit; Le, Yi; Kien Ng, Sook; Wolfgang, Christopher L.; Zellars, Richard C.; Ellsworth, Susannah G.; Ahuja, Nita

    2014-01-01

    High-dose-rate intraoperative radiation therapy (HDR-IORT) has historically provided effective local control (LC) for patients with unresectable and recurrent tumors. However, IORT is limited to only a few specialized institutions and it can be difficult to initiate an HDR-IORT program. Herein, we provide a brief overview on how to initiate and implement an HDR-IORT program for a selected group of patients with gastrointestinal and pelvic solid tumors using a multidisciplinary approach. Proper administration of HDR-IORT requires institutional support and a joint effort among physics staff, oncologists, surgeons, anesthesiologists, and nurses. In order to determine the true efficacy of IORT for various malignancies, collaboration among institutions with established IORT programs is needed. PMID:24790628

  19. Radical surgical resection and high-dose intraoperative radiation therapy (HDR-IORT) in patients with recurrent gynecologic cancers

    International Nuclear Information System (INIS)

    Gemignani, Mary L.; Alektiar, Kaled M.; Leitao, Mario; Mychalczak, Boris; Chi, Dennis; Venkatraman, Ennapadam; Barakat, Richard R.; Curtin, John P.

    2001-01-01

    Objective: To determine the outcome for patients with recurrent gynecologic tumors treated with radical resection and combined high-dose intraoperative radiation therapy (HDR-IORT). Methods and Materials: Between November 1993 and June 1998, 17 patients with recurrent gynecologic malignancies underwent radical surgical resection and high-dose-rate brachytherapy. The mean age of the study group was 49 years (range 28-72 years). The site of the primary tumor was the cervix in 9 (53%) patients, the uterus in 7 (41%) patients, and the vagina in 1 (6%) patient. The treatment for the primary disease was surgery with or without adjuvant radiation in 14 (82%) patients and definitive radiation in 3 (18%) patients. The current surgery consisted of exenterative surgery in 10 (59%) patients and tumor resection in 7 (41%) patients. Complete gross resection was achieved in 13 (76%) patients. The mean HDR-IORT dose was 14 Gy (range 12-15). Additional radiation in the form of permanent Iodine-125 implant was given to 3 of 4 patients with gross residual disease. The median peripheral dose was 140 Gy. Results: With a median follow-up of 20 months (range 3-65 months), the 3-year actuarial local control (LC) rate was 67%. In patients with complete gross resection, the 3-year LC rate was 83%, compared to 25% in patients with gross residual disease, p<0.01. The 3-year distant metastasis disease-free and overall survival rates were 54% and 54%, respectively. The complications were as follows: gastrointestinal obstruction, 4 (24%); wound complications, 4 (24%); abscesses, 3 (18%); peripheral neuropathy, 3 (18%); rectovaginal fistula, 2 (12%); and ureteral obstruction, 2 (12%). Conclusion: Radical surgical resection and combined IORT for patients with recurrent gynecologic tumors seems to provide a reasonable local-control rate in patients who have failed prior surgery and/or definitive radiation. Patient selection is very important, however, as only those patients with complete gross

  20. Post-operative high dose rate brachytherapy in patients with low to intermediate risk endometrial cancer

    International Nuclear Information System (INIS)

    Pearcey, R.G.; Petereit, D.G.

    2000-01-01

    This paper investigates the outcome using different dose/fractionation schedules in high dose rate (HDR) post-operative vaginal vault radiotherapy in patients with low to intermediate risk endometrial cancer. The world literature was reviewed and thirteen series were analyzed representing 1800 cases. A total of 12 vaginal vault recurrences were identified representing an overall vaginal control rate of 99.3%. A wide range of dose fractionation schedules and techniques have been reported. In order to analyze a dose response relationship for tumor control and complications, the biologically effective doses to the tumor and late responding tissues were calculated using the linear quadratic model. A threshold was identified for complications, but not vaginal control. While dose fractionation schedules that delivered a biologically effective dose to the late responding tissues in excess of 100 Gy 3 (LQED = 60 Gy) predicted for late complications, dose fractionation schedules that delivered a modest dose to the vaginal surface (50 Gy 10 or LQED = 30 Gy) appeared tumoricidal with vaginal control rates of at least 98%. By using convenient, modest dose fractionation schedules, HDR vaginal vault - brachytherapy yields very high local control and extremely low morbidity rates. (author)

  1. After low and high dose-rate interstitial brachytherapy followed by IMRT radiotherapy for intermediate and high risk prostate cancer

    International Nuclear Information System (INIS)

    Nakamura, Satoshi; Murakami, Naoya; Inaba, Koji; Wakita, Akihisa; Kobayashi, Kazuma; Takahashi, Kana; Okamoto, Hiroyuki; Umezawa, Rei; Morota, Madoka; Sumi, Minako; Igaki, Hiroshi; Ito, Yoshinori; Itami, Jun

    2016-01-01

    The study aimed to compare urinary symptoms in patients with clinically localized prostate cancer after a combination of either low-dose-rate or high-dose-rate interstitial brachytherapy along with intensity-modulated radiation therapy (LDR-ISBT + IMRT or HDR-ISBT + IMRT). From June 2009 to April 2014, 16 and 22 patients were treated with LDR-ISBT + IMRT and HDR-ISBT + IMRT, respectively. No patient from these groups was excluded from this study. The prescribed dose of LDR-ISBT, HDR-ISBT, and IMRT was 115 Gy, 20 Gy in 2 fractions, and 46 Gy in 23 fractions, respectively. Obstructive and irritative urinary symptoms were assessed by the International Prostate Symptom Score (IPSS) examined before and after treatments. After ISBT, IPSS was evaluated in the 1st and 4th weeks, then every 2–3 months for the 1st year, and every 6 months thereafter. The median follow-up of the patients treated with LDR-ISBT + IMRT and HDR-ISBT + IMRT was 1070.5 days and 1048.5 days, respectively (p = 0.321). The IPSS-increment in the LDR-ISBT + IMRT group was greater than that in the HDR-ISBT + IMRT between 91 and 180 days after ISBT (p = 0.015). In the LDR-ISBT + IMRT group, the IPSS took longer time to return to the initial level than in the HDR-ISBT + IMRT group (in LDR-ISBT + IMRT group, the recovery time was 90 days later). The dose to urethra showed a statistically significant association with the IPSS-increment in the irritative urinary symptoms (p = 0.011). Clinical outcomes were comparable between both the groups. Both therapeutic modalities are safe and well suited for patients with clinically localized prostate cancer; however, it took patients longer to recover from LDR-ISBT + IMRT than from HDR-ISBT + IMRT. It is possible that fast dose delivery induced early symptoms and early recovery, while gradual dose delivery induced late symptoms and late recovery. Urethral dose reductions were associated with small increments in IPSS

  2. Prediction of late rectal complication following high-dose-rate intracavitary brachytherapy in cancer of the uterine cervix

    International Nuclear Information System (INIS)

    Lee, Jeung Eun; Huh, Seung Jae; Park, Won; Lim, Do Hoon; Ahn, Yong Chan

    2003-01-01

    Although high-dose-rate intracavitary radiotherapy (HDR ICR) has been used in the treatment of cervical cancer, the potential for increased risk of late complication, most commonly in the rectum, is a major concern. We have previously reported on 136 patients treated with HDR brachytherapy between 1995 and 1999. The purpose of this study is to upgrade the previous data and confirm the correlation between late rectal complication and rectal dose in cervix cancer patients treated with HDR ICR. A retrospective analysis was performed for 222 patients with cervix cancer who were treated for curative intent with extemal beam radiotherapy (EBRT) and HDR ICR from July 1995 to December 2001. The median dose of EBRT was 50.4 (30.6-56.4) Gy with a daily fraction size 1.8 Gy. A total of six fractions of HDR ICR were given twice weekly with fraction size of 4 (3-5.5) Gy to A point by Iridium-192 source. The rectal dose was calculated at the rectal reference point using the barium contrast criteria in vivo measurement of the rectal dose was performed with thermoluminescent dosimeter (TLD) during HDR ICR. The median follow-up period was 39 months, ranging from 6 to 90 months. Twenty-one patients (9.5%) experienced late rectal bleeding, from 3 to 44 months (median, 13 months) after the completion of RT. The calculated rectal doses were not different between the patients with rectal bleeding and those without, but the measured rectal doses were higher in the complicated patients. The differences of the measured ICR rectal fractional dose, ICR total rectal dose, and total rectal biologically equivalent dose (BED) were statistically significant. When the measured ICR total rectal dose was beyond 16 Gy, when the ratio of the measured rectal dose to A point dose was beyond 70%, or when the measured rectal BED was over 110 GY 3 , a high possibility of late rectal complication was found. Late rectal complication was closely correlated with measured rectal dose by in vivo dosimetry using

  3. Comparison of radiation shielding requirements for HDR brachytherapy using 169Yb and 192Ir sources

    International Nuclear Information System (INIS)

    Lymperopoulou, G.; Papagiannis, P.; Sakelliou, L.; Georgiou, E.; Hourdakis, C. J.; Baltas, D.

    2006-01-01

    169 Yb has received a renewed focus lately as an alternative to 192 Ir sources for high dose rate (HDR) brachytherapy. Following the results of a recent work by our group which proved 169 Yb to be a good candidate for HDR prostate brachytherapy, this work seeks to quantify the radiation shielding requirements for 169 Yb HDR brachytherapy applications in comparison to the corresponding requirements for the current 192 Ir HDR brachytherapy standard. Monte Carlo simulation (MC) is used to obtain 169 Yb and 192 Ir broad beam transmission data through lead and concrete. Results are fitted to an analytical equation which can be used to readily calculate the barrier thickness required to achieve a given dose rate reduction. Shielding requirements for a HDR brachytherapy treatment room facility are presented as a function of distance, occupancy, dose limit, and facility workload, using analytical calculations for both 169 Yb and 192 Ir HDR sources. The barrier thickness required for 169 Yb is lower than that for 192 Ir by a factor of 4-5 for lead and 1.5-2 for concrete. Regarding 169 Yb HDR brachytherapy applications, the lead shielding requirements do not exceed 15 mm, even in highly conservative case scenarios. This allows for the construction of a lead door in most cases, thus avoiding the construction of a space consuming, specially designed maze. The effects of source structure, attenuation by the patient, and scatter conditions within an actual treatment room on the above-noted findings are also discussed using corresponding MC simulation results

  4. Postoperative vaginal cuff irradiation using high dose rate remote afterloading: a Phase II clinical protocol

    International Nuclear Information System (INIS)

    Noyes, William R.; Bastin, Kenneth; Edwards, Scott A.; Buchler, Dolores A.; Stitt, Judith A.; Thomadsen, Bruce R.; Fowler, Jack F.; Kinsella, Timothy J.

    1995-01-01

    Purpose: In September 1989, a postoperative Phase II high dose rate (HDR) brachytherapy protocol was started for International Federation of Gynecology and Obstetrics (FIGO) Stage I endometrial adenocarcinoma. This review reports the overall survival, local control, and complication rates for the initial 63 patients treated in this Phase II study. Methods and Materials: High dose rate brachytherapy was delivered using an Iridium-192 HDR remote afterloader. Sixty-three patients were entered into the Phase II protocol, each receiving two vaginal cuff treatments 1 week apart (range 4-12 days) with vaginal ovoids (diameter 2.0-3.0 cm). No patient received adjuvant external beam radiation. A dose of 32.4 Gy in two fractions was prescribed to the ovoid surface in 63 patients. The first three patients treated at our institution received 15, 16.2, and 29 Gy, respectively, to determine acute effects. Results: At a median follow-up of 1.6 years (range 0.75-4.3 years) no patient has developed a vaginal cuff recurrence. One regional recurrence (1.6%) occurred at 1.2 years at the pelvic side wall. This patient is alive and without evidence of disease 7 months after completion of salvage irradiation, which resulted in the only vaginal stenosis (1.6%). Fourteen patients (22%) experienced vaginal apex fibrosis by physical exam, which was clinically symptomatic in four patients. Two patients reported stress incontinence; however, these symptoms were noted prior to their HDR therapy. One patient died 2.4 years after HDR therapy due to cardiovascular disease without evidence of cancer at autopsy. Conclusion: Preliminary results of our phase II HDR vaginal cuff protocol for postoperative FIGO Stage IA, Grade 3 or Stage IB, Grade 1-2 patients demonstrate that 32.4 Gy in two fractions is well tolerated by the vaginal cuff mucosa. Local control appears comparable to our prior experience and others with low dose rate (LDR) brachytherapy. Additional patient accrual and further follow

  5. HDR brachytherapy in carcinoma of cervix: initial experience at AWARE hospitals

    International Nuclear Information System (INIS)

    Rajendran, M.; Reddy, K.D.; Reddy, R.M.; Reddy, J.M.; Reddy, B.V.N.; Kiran Kumar; Gopi, S.; Dharaniraj; Janardhanan

    2002-01-01

    High dose rate (HDR) brachytherapy is well established in the management of gynaecological malignancies. A report on the initial results of one and half year experience with a consistent dose/fractionation schedule and procedure of planning with delivery of treatment schedule is presented

  6. High-dose-rate brachytherapy using molds for oral cavity cancer. The technique and its limitations

    International Nuclear Information System (INIS)

    Nishimura, Yasumasa; Yokoe, Yoshihiko; Nagata, Yasushi; Okajima, Kaoru; Nishida, Mitsuo; Hiraoka, Masahiro

    1998-01-01

    With the availability of a high-dose-rate (HDR) remote afterloading device, a Phase I/II protocol was initiated at our institution to assess the toxicity and efficacy of HDR intracavitary brachytherapy, using molds, in the treatment of squamous cell carcinomas of the oral cavity. Eight patients with squamous cell carcinoma of the oral cavity were treated by the technique. The primary sites of the tumors were the buccal mucosa, oral floor, and gingiva. Two of the buccal mucosal cancers were located in the retromolar trigon. For each patient, a customized mold was fabricated, in which two to four afterloading catheters were placed for an 192 Ir HDR source. Four to seven fractions of 3-4 Gy, 5 mm below the mold surface, were given following external radiation therapy of 40-60 Gy/ 2 Gy. The total dose of HDR brachytherapy ranged from 16 to 28Gy. Although a good initial complete response rate of 7/8 (88%) was achieved, there was local recurrence in four of these seven patients. Both of the retromolar trigon tumors showed marginal recurrence. No serious (e.g., ulcer or bone exposure) late radiation damage has been observed thus far in the follow up period of 15-57 months. High-dose-rate brachytherapy using the mold technique seems a safe and useful method for selected early and superficial oral cavity cancer. However, it is not indicated for thick tumors and/or tumors located in the retromolar trigon. (author)

  7. SU-E-T-124: Dosimetric Comparison of HDR Brachytherapy and Intensity Modulated Proton Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Wu, J [Purdue University, West Lafayette, IN (United States); Wu, H [IUPUI, Indianapolis, IN (United States); Das, I [Indiana University- School of Medicine, Indianapolis, IN (United States)

    2014-06-01

    Purpose: Brachytherapy is known to be able to deliver more radiation dose to tumor while minimizing radiation dose to surrounding normal tissues. Proton therapy also provides superior dose distribution due to Bragg peak. Since both HDR and Intensity Modulated Proton Therapy (IMPT) are beneficial for their quick dose drop off, our goal in this study is to compare the pace of dose gradient drop-off between HDR and IMPT plans based on the same CT image data-set. In addition, normal tissues sparing were also compared among HDR, IMPT and SBRT. Methods: Five cervical cancer cases treated with EBRT + HDR boost combination with Tandem and Ovoid applicator were used for comparison purpose. Original HDR plans with prescribed dose of 5.5 Gy x 5 fractions were generated and optimized. The 100% isodose line of HDR plans was converted to a dose volume, and treated as CTV for IMPT and SBRT planning. The same HDR CT scans were also used for IMPT plan and SBRT plan for direct comparison. The philosophy of the IMPT and SBRT planning was to create the same CTV coverage as HDR plans. All three modalities treatment plans were compared to each other with a set of predetermined criteria. Results: With similar target volume coverage in cervix cancer boost treatment, HDR provides a slightly sharper dose drop-off from 100% to 50% isodose line, averagely in all directions compared to IMPT. However, IMPT demonstrated more dose gradient drop-off at the junction of the target and normal tissues by providing more normal tissue sparing and superior capability to reduce integral dose. Conclusion: IMPT is capable of providing comparable dose drop-off as HDR. IMPT can be explored as replacement for HDR brachytherapy in various applications.

  8. Dosimetric comparison between the microSelectron HDR 192Ir v2 source and the BEBIG 60Co source for HDR brachytherapy using the EGSnrc Monte Carlo transport code

    International Nuclear Information System (INIS)

    Anwarul Islam, M.; Akramuzzaman, M.M.; Zakaria, G.A.

    2012-01-01

    Manufacturing of miniaturized high activity 192 Ir sources have been made a market preference in modern brachytherapy. The smaller dimensions of the sources are flexible for smaller diameter of the applicators and it is also suitable for interstitial implants. Presently, miniaturized 60 Co HDR sources have been made available with identical dimensions to those of 192 Ir sources. 60 Co sources have an advantage of longer half life while comparing with 192 Ir source. High dose rate brachytherapy sources with longer half life are logically pragmatic solution for developing country in economic point of view. This study is aimed to compare the TG-43U1 dosimetric parameters for new BEBIG 60 Co HDR and new microSelectron 192 Ir HDR sources. Dosimetric parameters are calculated using EGSnrc-based Monte Carlo simulation code accordance with the AAPM TG-43 formalism for microSelectron HDR 192 Ir v2 and new BEBIG 60 Co HDR sources. Air-kerma strength per unit source activity, calculated in dry air are 9.698x10 -8 ± 0.55% U Bq -1 and 3.039x10 -7 ± 0.41% U Bq -1 for the above mentioned two sources, respectively. The calculated dose rate constants per unit air-kerma strength in water medium are 1.116±0.12% cGy h -1 U -1 and 1.097±0.12% cGy h -1 U -1 , respectively, for the two sources. The values of radial dose function for distances up to 1 cm and more than 22 cm for BEBIG 60 Co HDR source are higher than that of other source. The anisotropic values are sharply increased to the longitudinal sides of the BEBIG 60 Co source and the rise is comparatively sharper than that of the other source. Tissue dependence of the absorbed dose has been investigated with vacuum phantom for breast, compact bone, blood, lung, thyroid, soft tissue, testis, and muscle. No significant variation is noted at 5 cm of radial distance in this regard while comparing the two sources except for lung tissues. The true dose rates are calculated with considering photon as well as electron transport using

  9. Bronchial mucosal tolerance following high dose rate endobronchial brachytherapy: clinical and laboratory correlates in late complication assessment of fatal hemoptysis

    International Nuclear Information System (INIS)

    Chen, Y.Y.; McDonald, S.; Nakamura, C.; Philips, A.; Ojomo, Karanita; Hernady, E.; Williams, J. P.; Smudzin, T.; Johnstone, D.; Feins, R.; Speiser, B.L.; Rubin, P.

    1996-01-01

    Purpose: This study reviews treatment related late complications following high dose rate endobronchial brachytherapy (HDR-EB) for primary lung cancers. Radiation dose contribution from HDR-EB treatment alone, or combined HDR-EB and external beam (EBRT) were analyzed in relation to the linear representation of the tracheobronchial anatomy. Results were presented in a dose volume histogram (DVH) analysis for the risk estimate of late complications from HDR-EB. Bronchial mucosal tolerance is estimated from the clinical experience study and histopathologic changes in laboratory animals treated with HDR-EB. Methods: 1.) There were forty one patients with primary lung cancer received HDR-EB as part of the radiation treatment between December 1990 and June 1994. Six of these developed late complications manifested as either fatal hemoptysis or endobronchial deposition of fibrinous material/bronchial stenosis. Treatment planning films were reviewed to map the volume treated with HDR-EB and EBRT along the tracheobronchial segments. DVH was constructed and compared for patients with and without late complications. Other clinical parameters of interest which were analyzed included: dose per fraction, EBRT total dose, HDR total dose, combined EBRT and HDR total dose, number of catheters per treatment, and points of prescriptions for HDR-EB. 2.) Forty four New Zealand White rabbits underwent HDR-EB of the major airway to a treatment length of 2 cm (1 cm above and below the carina) to a single fraction dose of 10 Gy, 30 Gy, or 50 Gy. Histopathologic changes were examined at 7, 14, 28, and 56 days post-treatment and compared with the control rabbits which received no irradiation. Results: 1.) The late complication rate is 14.5% with three patients developing fibrinous deposits/bronchial stenosis and four patients who experienced fatal hemoptysis in a total of six patients. There is a significant difference in DVH of HDR-EB treatment in the tracheobronchial high dose region

  10. Modulation of toxicity following external beam irradiation preceded by high-dose rate brachytherapy in inoperable oesophageal cancer

    Energy Technology Data Exchange (ETDEWEB)

    Taal, B.G.; Aleman, B.M.P.; Koning, C.C.E.; Boot, H. [Nederlands Kanker Inst. `Antoni van Leeuwenhoekhuis`, Amsterdam (Netherlands)

    1996-09-01

    To induce fast relief of dysphagia in inoperable oesephageal cancer, we applied high-dose rate (HDR) intraluminal irradiation followed by external irradiation (EBRT) in a phase II study. 15 patients (group A: n = 15; 10 men, 5 women; median age 66 years) were treated with 10 Gy HDR brachytherapy plus 40 Gy EBRT (15 fractions of 2.67 Gy). Severe side-effects were encountered in 60% of patients: 3 late ulceration, 2 pending fistula and 2 patients with fatal haemorrhage after an interval of 6 months. Overall response was excellent: 9 complete remissions (60%) and 6 partial responses (40%). Because of the high toxicity rate, in a subsequent study (group B: n = 30; 23 mean, 7 women; median age 66 years) the EBRT scheme was changed using smaller fractions (2.0 Gy) to reach the same total dose of 40 Gy. The complication rate (17%) was significantly reduced, while the overall response remained excellent (83%): 17 complete and 8 partial responses. The impressive change in complication rate of HDR brachytherapy and EBRT stresses the impact of the fraction per dose and illustrates the small therapeutic margins. (author).

  11. Modulation of toxicity following external beam irradiation preceded by high-dose rate brachytherapy in inoperable oesophageal cancer

    International Nuclear Information System (INIS)

    Taal, B.G.; Aleman, B.M.P.; Koning, C.C.E.; Boot, H.

    1996-01-01

    To induce fast relief of dysphagia in inoperable oesephageal cancer, we applied high-dose rate (HDR) intraluminal irradiation followed by external irradiation (EBRT) in a phase II study. 15 patients (group A: n = 15; 10 men, 5 women; median age 66 years) were treated with 10 Gy HDR brachytherapy plus 40 Gy EBRT (15 fractions of 2.67 Gy). Severe side-effects were encountered in 60% of patients: 3 late ulceration, 2 pending fistula and 2 patients with fatal haemorrhage after an interval of 6 months. Overall response was excellent: 9 complete remissions (60%) and 6 partial responses (40%). Because of the high toxicity rate, in a subsequent study (group B: n = 30; 23 mean, 7 women; median age 66 years) the EBRT scheme was changed using smaller fractions (2.0 Gy) to reach the same total dose of 40 Gy. The complication rate (17%) was significantly reduced, while the overall response remained excellent (83%): 17 complete and 8 partial responses. The impressive change in complication rate of HDR brachytherapy and EBRT stresses the impact of the fraction per dose and illustrates the small therapeutic margins. (author)

  12. A gEUD-based inverse planning technique for HDR prostate brachytherapy: Feasibility study

    Energy Technology Data Exchange (ETDEWEB)

    Giantsoudi, D. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78229 (United States); Department of Radiation Oncology, Francis H. Burr Proton Therapy Center, Boston, Massachusetts 02114 (United States); Baltas, D. [Department of Medical Physics and Engineering, Strahlenklinik, Klinikum Offenbach GmbH, 63069 Offenbach (Germany); Nuclear and Particle Physics Section, Physics Department, University of Athens, 15701 Athens (Greece); Karabis, A. [Pi-Medical Ltd., Athens 10676 (Greece); Mavroidis, P. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78299 and Department of Medical Radiation Physics, Karolinska Institutet and Stockholm University, 17176 (Sweden); Zamboglou, N.; Tselis, N. [Strahlenklinik, Klinikum Offenbach GmbH, 63069 Offenbach (Germany); Shi, C. [St. Vincent' s Medical Center, 2800 Main Street, Bridgeport, Connecticut 06606 (United States); Papanikolaou, N. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78299 (United States)

    2013-04-15

    Purpose: The purpose of this work was to study the feasibility of a new inverse planning technique based on the generalized equivalent uniform dose for image-guided high dose rate (HDR) prostate cancer brachytherapy in comparison to conventional dose-volume based optimization. Methods: The quality of 12 clinical HDR brachytherapy implants for prostate utilizing HIPO (Hybrid Inverse Planning Optimization) is compared with alternative plans, which were produced through inverse planning using the generalized equivalent uniform dose (gEUD). All the common dose-volume indices for the prostate and the organs at risk were considered together with radiobiological measures. The clinical effectiveness of the different dose distributions was investigated by comparing dose volume histogram and gEUD evaluators. Results: Our results demonstrate the feasibility of gEUD-based inverse planning in HDR brachytherapy implants for prostate. A statistically significant decrease in D{sub 10} or/and final gEUD values for the organs at risk (urethra, bladder, and rectum) was found while improving dose homogeneity or dose conformity of the target volume. Conclusions: Following the promising results of gEUD-based optimization in intensity modulated radiation therapy treatment optimization, as reported in the literature, the implementation of a similar model in HDR brachytherapy treatment plan optimization is suggested by this study. The potential of improved sparing of organs at risk was shown for various gEUD-based optimization parameter protocols, which indicates the ability of this method to adapt to the user's preferences.

  13. NOTE: Monte Carlo evaluation of kerma in an HDR brachytherapy bunker

    Science.gov (United States)

    Pérez-Calatayud, J.; Granero, D.; Ballester, F.; Casal, E.; Crispin, V.; Puchades, V.; León, A.; Verdú, G.

    2004-12-01

    In recent years, the use of high dose rate (HDR) after-loader machines has greatly increased due to the shift from traditional Cs-137/Ir-192 low dose rate (LDR) to HDR brachytherapy. The method used to calculate the required concrete and, where appropriate, lead shielding in the door is based on analytical methods provided by documents published by the ICRP, the IAEA and the NCRP. The purpose of this study is to perform a more realistic kerma evaluation at the entrance maze door of an HDR bunker using the Monte Carlo code GEANT4. The Monte Carlo results were validated experimentally. The spectrum at the maze entrance door, obtained with Monte Carlo, has an average energy of about 110 keV, maintaining a similar value along the length of the maze. The comparison of results from the aforementioned values with the Monte Carlo ones shows that results obtained using the albedo coefficient from the ICRP document more closely match those given by the Monte Carlo method, although the maximum value given by MC calculations is 30% greater.

  14. High-Dose-Rate Monotherapy: Safe and Effective Brachytherapy for Patients With Localized Prostate Cancer

    International Nuclear Information System (INIS)

    Demanes, D. Jeffrey; Martinez, Alvaro A.; Ghilezan, Michel; Hill, Dennis R.; Schour, Lionel; Brandt, David; Gustafson, Gary

    2011-01-01

    Purpose: High-dose-rate (HDR) brachytherapy used as the only treatment (monotherapy) for early prostate cancer is consistent with current concepts in prostate radiobiology, and the dose is reliably delivered in a prospectively defined anatomic distribution that meets all the requirements for safe and effective therapy. We report the disease control and toxicity of HDR monotherapy from California Endocurietherapy (CET) and William Beaumont Hospital (WBH) in low- and intermediate-risk prostate cancer patients. Methods and Materials: There were 298 patients with localized prostate cancer treated with HDR monotherapy between 1996 and 2005. Two biologically equivalent hypofractionation protocols were used. At CET the dose was 42 Gy in six fractions (two implantations 1 week apart) delivered to a computed tomography–defined planning treatment volume. At WBH the dose was 38 Gy in four fractions (one implantation) based on intraoperative transrectal ultrasound real-time treatment planning. The bladder, urethral, and rectal dose constraints were similar. Toxicity was scored with the National Cancer Institute Common Toxicity Criteria for Adverse Events version 3. Results: The median follow-up time was 5.2 years. The median age of the patients was 63 years, and the median value of the pretreatment prostate-specific antigen was 6.0 ng/mL. The 8-year results were 99% local control, 97% biochemical control (nadir +2), 99% distant metastasis–free survival, 99% cause-specific survival, and 95% overall survival. Toxicity was scored per event, meaning that an individual patient with more than one symptom was represented repeatedly in the morbidity data table. Genitourinary toxicity consisted of 10% transient Grade 2 urinary frequency or urgency and 3% Grade 3 episode of urinary retention. Gastrointestinal toxicity was <1%. Conclusions: High disease control rates and low morbidity demonstrate that HDR monotherapy is safe and effective for patients with localized prostate cancer.

  15. Post operative high dose rate intravaginal irradiation in endometrial cancer: a safe and effective outpatient treatment

    International Nuclear Information System (INIS)

    Chen, Peter; Gibbons, Susan; Vicini, Frank; Weiner, Sheldon; Dmuchowski, Carl; Mele, Beth; Brabbins, Donald; Jennings, John; Gustafson, Gary; Martinez, Alvaro

    1995-01-01

    Purpose: We reviewed our experience with out patient high dose rate (HDR) intravaginal irradiation given post-operatively in endometrial cancer to assess local control, survival, and toxicity when used alone or in combination with external beam irradiation. Methods and Materials: From (12(88)) to (12(92)), 78 patients underwent TAH/BSO and received post-operative HDR intravaginal irradiation for endometrial cancer. Pathologic stage distribution was IB/IC: 56%, II: 22%, III: 22%. Adjuvant therapy was given in one of three schemes: HDR vaginal radiation alone (6 weekly fractions of 500 cGy prescribed 5 mm from the applicator surface treating the upper 4 cm of the vagina), pelvic irradiation with vaginal HDR (500 cGy x 4 weekly fractions) or whole abdomen/pelvic irradiation (WAPI) with vaginal HDR treatment (500 cGy x 3 weekly fractions). Prior to the first HDR vaginal treatment, a simulation with placement of vaginal apex metallic markers was performed to assure proper positioning of the intravaginal cylinders. Pelvic midline blocking was designed from the HDR intravaginal simulation films. The 55 patients who underwent combined external beam irradiation/brachytherapy received a median dose to the pelvis of 5040 cGy (range 25.2-51.6 Gy), and a median total vaginal dose of 5060 cGy (range 30.0-57.6 Gy). Results: Median follow-up is 37 months (range 6-73 months). Local control (vaginally) is 98.7%. The one vaginal failure was in the distal vagina, outside the treatment volume. All other failures (4) were distant with the vagina controlled [3 intra-abdominal and one bone/intra-abdominal]. For stages I and II, the disease free survival is 92.8%. For stage III the disease free survival is 86.5%. Median overall time to failure is 14.3 months (range 8.5-18.6 months). In terms of acute toxicity, no grade 3-4 acute toxicity of the vagina or bladder was seen. However, 9% acute GI toxicity was encountered. Chronic grade 1-2 toxicities included: vaginal 21.8% (foreshortening and

  16. SU-E-J-270: Study of PET Response to HDR Brachytherapy of Rectal Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Hobbs, R; Le, Y; Armour, E; Efron, J; Azad, N; Wahl, R; Gearhart, S; Herman, J [Johns Hopkins University, Baltimore, MD (United States)

    2014-06-01

    Purpose: Dose-response studies in radiation therapy are typically using single response values for tumors across ensembles of tumors. Using the high dose rate (HDR) treatment plan dose grid and pre- and post-therapy FDG-PET images, we look for correlations between voxelized dose and FDG uptake response in individual tumors. Methods: Fifteen patients were treated for localized rectal cancer using 192Ir HDR brachytherapy in conjunction with surgery. FDG-PET images were acquired before HDR therapy and 6–8 weeks after treatment (prior to surgery). Treatment planning was done on a commercial workstation and the dose grid was calculated. The two PETs and the treatment dose grid were registered to each other using non-rigid registration. The difference in PET SUV values before and after HDR was plotted versus absorbed radiation dose for each voxel. The voxels were then separated into bins for every 400 cGy of absorbed dose and the bin average values plotted similarly. Results: Individual voxel doses did not correlate with PET response; however, when group into tumor subregions corresponding to dose bins, eighty percent of the patients showed a significant positive correlation (R2 > 0) between PET uptake difference in the targeted region and the absorbed dose. Conclusion: By considering larger ensembles of voxels, such as organ average absorbed dose or the dose bins considered here, valuable information may be obtained. The dose-response correlations as measured by FDG-PET difference potentially underlines the importance of FDG-PET as a measure of response, as well as the value of voxelized information.

  17. Time to PSA rise differentiates the PSA bounce after HDR and LDR brachytherapy of prostate cancer.

    Science.gov (United States)

    Burchardt, Wojciech; Skowronek, Janusz

    2018-02-01

    To investigate the differences in prostate-specific antigen (PSA) bounce (PB) after high-dose-rate (HDR-BT) or low-dose-rate (LDR-BT) brachytherapy alone in prostate cancer patients. Ninety-four patients with localized prostate cancer (T1-T2cN0), age ranged 50-81 years, were treated with brachytherapy alone between 2008 and 2010. Patients were diagnosed with adenocarcinoma, Gleason score ≤ 7. The LDR-BT total dose was 144-145 Gy, in HDR-BT - 3 fractions of 10.5 or 15 Gy. The initial PSA level (iPSA) was assessed before treatment, then PSA was rated every 3 months over the first 2 years, and every 6 months during the next 3 years. Median follow-up was 3.0 years. Mean iPSA was 7.8 ng/ml. In 58 cases, PSA decreased gradually without PB or biochemical failure (BF). In 24% of patients, PB was observed. In 23 cases (24%), PB was observed using 0.2 ng/ml definition; in 10 cases (11%), BF was diagnosed using nadir + 2 ng/ml definition. The HDR-BT and LDR-BT techniques were not associated with higher level of PB (26 vs. 22%, p = 0.497). Time to the first PSA rise finished with PB was significantly shorter after HDR-BT then after LDR-BT (median, 10.5 vs. 18.0 months) during follow-up. Predictors for PB were observed only after HDR-BT. Androgen deprivation therapy (ADT) and higher Gleason score decreased the risk of PB (HR = 0.11, p = 0.03; HR = 0.51, p = 0.01). The higher PSA nadir and longer time to PSA nadir increased the risk of PB (HR 3.46, p = 0.02; HR 1.04, p = 0.04). There was no predictors for PB after LDR-BT. HDR-BT and LDR-BT for low and intermediate risk prostate cancer had similar PB rate. The PB occurred earlier after HDR-BT than after LDR-BT. ADT and higher Gleason score decreased, and higher PSA nadir and longer time to PSA nadir increased the risk of PB after HDR-BT.

  18. Current situation of high-dose-rate brachytherapy for cervical cancer in Brazil

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Rogerio Matias Vidal da; Souza, Divanizia do Nascimento, E-mail: rmv.fisica@gmail.com [Universidade Federal de Sergipe (UFS), Sao Cristovao, SE (Brazil); Pinezi, Juliana Castro Dourado [Pontificia Universidade Catolica de Goias (PUC-Goias), Goiania, GO (Brazil); Macedo, Luiz Eduardo Andrade [Hospital Chama, Arapiraca, AL (Brazil)

    2014-05-15

    To assess the current situation of high-dose-rate (HDR) brachytherapy for cancer of the cervix in Brazil, regarding apparatuses, planning methods, prescription, fractionation schedule and evaluation of dose in organs at risk. Materials and methods: in the period between March/2012 and May/2013, a multiple choice questionnaire was developed and sent to 89 Brazilian hospitals which perform HDR brachytherapy. Results: sixty-one services answered the questionnaire. All regions of the country experienced a sharp increase in the number of HDR brachytherapy services in the period from 2001 to 2013. As regards planning, although a three-dimensional planning software was available in 91% of the centers, conventional radiography was mentioned by 92% of the respondents as their routine imaging method for such a purpose. Approximately 35% of respondents said that brachytherapy sessions are performed after teletherapy. The scheme of four 7 Gy intracavitary insertions was mentioned as the most frequently practiced. Conclusion: the authors observed that professionals have difficulty accessing adjuvant three-dimensional planning tools such as computed tomography and magnetic resonance imaging. (author)

  19. Dose rate and fractionation: Relative importance in radiation for bone marrow transplantation

    International Nuclear Information System (INIS)

    Tarbell, N.J.; Rosenblatt, M.; Mauch, P.; Hellman, S.

    1987-01-01

    The optimal dose rate and fractionation schedules for total body irradiation (TBI) in bone marrow transplantation (BMT) are presently unknown. This study compares several fractionation and dose rate schedules that are currently in clinical use. C/sub 3/H/HeJ were given TBI and the bone marrow survival fraction was calculated using the CFU's assay. Irradiation was given as low dose rate (LDR) at 5 cGy/min or high dose rate (HDR) at 80 cGy/min, in single fraction (SF) and fractionated (FX) regimens. These results indicate no increase in survival for the normal bone marrow stem cells with fractionation either at high or low dose-rates. In fact, fractionation seemed to decrease the bone marrow survival over single fraction radiation

  20. Genetic effect of low dose rate radiation on human cells immortalized with the hTERT gene

    International Nuclear Information System (INIS)

    Nakamura, Hideaki; Fukami, Hiroko; Hayashi, Yuko; Kiyono, Tohru; Ishizaki, Kanji; Tachibana, Akira; Nakatsugawa, Shigekazu; Hamaguchi, Michinari

    2003-01-01

    We established immortal human cells by introducing the hTERT gene into skin fibroblast cells obtained from normal (SuSa) and ataxia telangiectasia (AT: AT1OS) individuals of Japanese origin. These immortalized cells showed the same characteristics as the original cells except expanded life span. We irradiated SuSa/T-n and AT1OS/T-n cells with low-dose-rate (LDR; 0.3 mGy/min) irradiation at confluent state in low-serum medium. Then, survival rate and micronucleus frequency of each cell line were analyzed. In SuSa/T-n cells, frequency of HPRT mutation induction was also determined by 6TG selection. In SuSa/T-n cells, survival rate and micronucleus frequency showed higher resistance after irradiation with LDR than high-dose-rate (HDR; 2 Gy/min) irradiation. In contrast, no significant difference was observed in survival and micronucleus induction in AT1OS/T-n cells between HDR and LDR irradiation, suggesting that AT1OS/T-n cells may have some defect in DNA repair activity. In SuSa/T-n cells, the frequency of HPRT mutation after LDR irradiation decreased to approximately one eighth that after HDR irradiation. (author)

  1. Dosimetric advancement of high-dose-rate after-loading 192Ir source

    International Nuclear Information System (INIS)

    Zhang Shuxu; Li Wenhua; Xu Hairong

    2004-01-01

    High-dose-rate (HDR) 192 Ir source is a nuclide commonly used in the brachytherapy system. The basic dosimetry data of the near source area is usually measured by pin ion chambers or TLD techniques, but these methods have a lower spatial resolution than Electron spin resonance (ESR) dosimetry which has a spatial resolution of 156 μm, and the Monte Carlo photon transport simulations are taken as the golden standard of those measures. The precision in two-dimensional dose distribution measured by GafChromic film is reported to be 1.0%. In vivo dosimetry using TLD during HDR intracavitary after-loading brachytherapy is a good predictor of late rectal complications. The accuracy of magnetic resonance imaging (MRI) Fricke-gel dosimetry for three-dimensional dose distribution is about 2.5% with a spatial resolution of 1.56 mm. The optical computed tomography polymer gel dosimetry has a unique advance than MRI gel dosimetry

  2. Increasing Fractional Doses Increases the Probability of Benign PSA Bounce in Patients Undergoing Definitive HDR Brachytherapy for Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Hauck, Carlin R.; Ye, Hong; Chen, Peter Y.; Gustafson, Gary S.; Limbacher, Amy; Krauss, Daniel J., E-mail: Daniel.krauss@beaumont.edu

    2017-05-01

    Purpose: Prostate-specific antigen (PSA) bounce is a temporary elevation of the PSA level above a prior nadir. The purpose of this study was to determine whether the frequency of a PSA bounce following high-dose-rate (HDR) interstitial brachytherapy for the treatment of prostate cancer is associated with individual treatment fraction size. Methods and Materials: Between 1999 and 2014, 554 patients underwent treatment of low- or intermediate-risk prostate cancer with definitive HDR brachytherapy as monotherapy and had ≥3 subsequent PSA measurements. Four different fraction sizes were used: 950 cGy × 4 fractions, 1200 cGy × 2 fractions, 1350 cGy × 2 fractions, 1900 cGy × 1 fraction. Four definitions of PSA bounce were applied: ≥0.2, ≥0.5, ≥1.0, and ≥2.0 ng/mL above the prior nadir with a subsequent return to the nadir. Results: The median follow-up period was 3.7 years. The actuarial 3-year rate of PSA bounce for the entire cohort was 41.3%, 28.4%, 17.4%, and 6.8% for nadir +0.2, +0.5, +1.0, and +2.0 ng/mL, respectively. The 3-year rate of PSA bounce >0.2 ng/mL was 42.2%, 32.1%, 41.0%, and 59.1% for the 950-, 1200-, 1350-, and 1900-cGy/fraction levels, respectively (P=.002). The hazard ratio for bounce >0.2 ng/mL for patients receiving a single fraction of 1900 cGy compared with those receiving treatment in multiple fractions was 1.786 (P=.024). For patients treated with a single 1900-cGy fraction, the 1-, 2-, and 3-year rates of PSA bounce exceeding the Phoenix biochemical failure definition (nadir +2 ng/mL) were 4.5%, 18.7%, and 18.7%, respectively, higher than the rates for all other administered dose levels (P=.025). Conclusions: The incidence of PSA bounce increases with single-fraction HDR treatment. Knowledge of posttreatment PSA kinetics may aid in decision making regarding management of potential biochemical failures.

  3. Increasing Fractional Doses Increases the Probability of Benign PSA Bounce in Patients Undergoing Definitive HDR Brachytherapy for Prostate Cancer

    International Nuclear Information System (INIS)

    Hauck, Carlin R.; Ye, Hong; Chen, Peter Y.; Gustafson, Gary S.; Limbacher, Amy; Krauss, Daniel J.

    2017-01-01

    Purpose: Prostate-specific antigen (PSA) bounce is a temporary elevation of the PSA level above a prior nadir. The purpose of this study was to determine whether the frequency of a PSA bounce following high-dose-rate (HDR) interstitial brachytherapy for the treatment of prostate cancer is associated with individual treatment fraction size. Methods and Materials: Between 1999 and 2014, 554 patients underwent treatment of low- or intermediate-risk prostate cancer with definitive HDR brachytherapy as monotherapy and had ≥3 subsequent PSA measurements. Four different fraction sizes were used: 950 cGy × 4 fractions, 1200 cGy × 2 fractions, 1350 cGy × 2 fractions, 1900 cGy × 1 fraction. Four definitions of PSA bounce were applied: ≥0.2, ≥0.5, ≥1.0, and ≥2.0 ng/mL above the prior nadir with a subsequent return to the nadir. Results: The median follow-up period was 3.7 years. The actuarial 3-year rate of PSA bounce for the entire cohort was 41.3%, 28.4%, 17.4%, and 6.8% for nadir +0.2, +0.5, +1.0, and +2.0 ng/mL, respectively. The 3-year rate of PSA bounce >0.2 ng/mL was 42.2%, 32.1%, 41.0%, and 59.1% for the 950-, 1200-, 1350-, and 1900-cGy/fraction levels, respectively (P=.002). The hazard ratio for bounce >0.2 ng/mL for patients receiving a single fraction of 1900 cGy compared with those receiving treatment in multiple fractions was 1.786 (P=.024). For patients treated with a single 1900-cGy fraction, the 1-, 2-, and 3-year rates of PSA bounce exceeding the Phoenix biochemical failure definition (nadir +2 ng/mL) were 4.5%, 18.7%, and 18.7%, respectively, higher than the rates for all other administered dose levels (P=.025). Conclusions: The incidence of PSA bounce increases with single-fraction HDR treatment. Knowledge of posttreatment PSA kinetics may aid in decision making regarding management of potential biochemical failures.

  4. Verification of the plan dosimetry for high dose rate brachytherapy using metal-oxide-semiconductor field effect transistor detectors

    International Nuclear Information System (INIS)

    Qi Zhenyu; Deng Xiaowu; Huang Shaomin; Lu Jie; Lerch, Michael; Cutajar, Dean; Rosenfeld, Anatoly

    2007-01-01

    The feasibility of a recently designed metal-oxide-semiconductor field effect transistor (MOSFET) dosimetry system for dose verification of high dose rate (HDR) brachytherapy treatment planning was investigated. MOSFET detectors were calibrated with a 0.6 cm 3 NE-2571 Farmer-type ionization chamber in water. Key characteristics of the MOSFET detectors, such as the energy dependence, that will affect phantom measurements with HDR 192 Ir sources were measured. The MOSFET detector was then applied to verify the dosimetric accuracy of HDR brachytherapy treatments in a custom-made water phantom. Three MOSFET detectors were calibrated independently, with the calibration factors ranging from 0.187 to 0.215 cGy/mV. A distance dependent energy response was observed, significant within 2 cm from the source. The new MOSFET detector has a good reproducibility ( 2 =1). It was observed that the MOSFET detectors had a linear response to dose until the threshold voltage reached approximately 24 V for 192 Ir source measurements. Further comparison of phantom measurements using MOSFET detectors with dose calculations by a commercial treatment planning system for computed tomography-based brachytherapy treatment plans showed that the mean relative deviation was 2.2±0.2% for dose points 1 cm away from the source and 2.0±0.1% for dose points located 2 cm away. The percentage deviations between the measured doses and the planned doses were below 5% for all the measurements. The MOSFET detector, with its advantages of small physical size and ease of use, is a reliable tool for quality assurance of HDR brachytherapy. The phantom verification method described here is universal and can be applied to other HDR brachytherapy treatments

  5. Interim report of image-guided conformal high-dose-rate brachytherapy for patients with unfavorable prostate cancer: the William Beaumont Phase II dose-escalating trial

    International Nuclear Information System (INIS)

    Martinez, Alvaro A.; Kestin, Larry L.; Stromberg, Jannifer S.; Gonzalez, Jose A.; Wallace, Michelle; Gustafson, Gary S.; Edmundson, Gregory K.; Spencer, William; Vicini, Frank A.

    2000-01-01

    Purpose: We analyzed our institution's experience treating patients with unfavorable prostate cancer in a prospective Phase II dose-escalating trial of external beam radiation therapy (EBRT) integrated with conformal high-dose-rate (HDR) brachytherapy boosts. This interim report discusses treatment outcome and prognostic factors using this treatment approach. Methods and Materials: From November 1991 through February 1998, 142 patients with unfavorable prostate cancer were prospectively treated in a dose-escalating trial with pelvic EBRT in combination with outpatient HDR brachytherapy at William Beaumont Hospital. Patients with any of the following characteristics were eligible: pretreatment prostate-specific antigen (PSA) ≥ 10.0 ng/ml, Gleason score ≥ 7, or clinical stage T2b or higher. All patients received pelvic EBRT to a median total dose of 46.0 Gy. Pelvic EBRT was integrated with ultrasound-guided transperineal conformal interstitial iridium-192 HDR implants. From 1991 to 1995, 58 patients underwent three conformal interstitial HDR implants during the first, second, and third weeks of pelvic EBRT. After October 1995, 84 patients received two interstitial implants during the first and third weeks of pelvic EBRT. The dose delivered via interstitial brachytherapy was escalated from 5.50 Gy to 6.50 Gy for each implant in those patients receiving three implants, and subsequently, from 8.25 Gy to 9.50 Gy per fraction in those patients receiving two implants. To improve implant quality and reduce operator dependency, an on-line, image-guided interactive dose optimization program was utilized during each HDR implant. No patient received hormonal therapy unless treatment failure was documented. The median follow-up was 2.1 years (range: 0.2-7.2 years). Biochemical failure was defined according to the American Society for Therapeutic Radiology and Oncology Consensus Panel definition. Results: The pretreatment PSA level was ≥ 10.0 ng/ml in 51% of patients. The

  6. MR-based source localization for MR-guided HDR brachytherapy

    Science.gov (United States)

    Beld, E.; Moerland, M. A.; Zijlstra, F.; Viergever, M. A.; Lagendijk, J. J. W.; Seevinck, P. R.

    2018-04-01

    For the purpose of MR-guided high-dose-rate (HDR) brachytherapy, a method for real-time localization of an HDR brachytherapy source was developed, which requires high spatial and temporal resolutions. MR-based localization of an HDR source serves two main aims. First, it enables real-time treatment verification by determination of the HDR source positions during treatment. Second, when using a dummy source, MR-based source localization provides an automatic detection of the source dwell positions after catheter insertion, allowing elimination of the catheter reconstruction procedure. Localization of the HDR source was conducted by simulation of the MR artifacts, followed by a phase correlation localization algorithm applied to the MR images and the simulated images, to determine the position of the HDR source in the MR images. To increase the temporal resolution of the MR acquisition, the spatial resolution was decreased, and a subpixel localization operation was introduced. Furthermore, parallel imaging (sensitivity encoding) was applied to further decrease the MR scan time. The localization method was validated by a comparison with CT, and the accuracy and precision were investigated. The results demonstrated that the described method could be used to determine the HDR source position with a high accuracy (0.4–0.6 mm) and a high precision (⩽0.1 mm), at high temporal resolutions (0.15–1.2 s per slice). This would enable real-time treatment verification as well as an automatic detection of the source dwell positions.

  7. Validating Fricke dosimetry for the measurement of absorbed dose to water for HDR 192Ir brachytherapy: a comparison between primary standards of the LCR, Brazil, and the NRC, Canada

    Science.gov (United States)

    Salata, Camila; Gazineu David, Mariano; de Almeida, Carlos Eduardo; El Gamal, Islam; Cojocaru, Claudiu; Mainegra-Hing, Ernesto; McEwen, Malcom

    2018-04-01

    Two Fricke-based absorbed dose to water standards for HDR Ir-192 dosimetry, developed independently by the LCR in Brazil and the NRC in Canada have been compared. The agreement in the determination of the dose rate from a HDR Ir-192 source at 1 cm in a water phantom was found to be within the k  =  1 combined measurement uncertainties of the two standards: D NRC/D LCR  =  1.011, standard uncertainty  =  2.2%. The dose-based standards also agreed within the uncertainties with the manufacturer’s stated dose rate value, which is traceable to a national standard of air kerma. A number of possible influence quantities were investigated, including the specific method for producing the ferrous-sulphate Fricke solution, the geometry of the holder, and the Monte Carlo code used to determine correction factors. The comparison highlighted the lack of data on the determination of G(Fe3+) in this energy range and the possibilities for further development of the holders used to contain the Fricke solution. The comparison also confirmed the suitability of Fricke dosimetry for Ir-192 primary standard dose rate determinations at therapy dose levels.

  8. Dosimetric evaluation of a novel high dose rate (HDR) intraluminal / interstitial brachytherapy applicator for gastrointestinal and bladder cancers

    Science.gov (United States)

    Aghamiri, Seyyed Mahmoud Reza; Najarian, Siamak; Jaberi, Ramin

    2010-01-01

    High dose rate (HDR) brachytherapy is one of the accepted treatment modalities in gastroâ€intestinal tract and bladder carcinomas. Considering the shortcoming of contact brachytherapy routinely used in gastrointestinal tract in treatment of big tumors or invasive method of bladder treatment, an intraluminal applicator with the capability of insertion into the tumor depth seems to be useful. This study presents some dosimetric evaluations to introduce this applicator to the clinical use. The radiation attenuation characteristics of the applicator were evaluated by means of two dosimetric methods including wellâ€type chamber and radiochromic film. The proposed 110 cm long applicator has a flexible structure made of stainless steel for easy passage through lumens and a needle tip to drill into big tumors. The 2 mm diameter of the applicator is thick enough for source transition, while easy passage through any narrow lumen such as endoscope or cystoscope working channel is ensured. Wellâ€chamber results showed an acceptably low attenuation of this steel springy applicator. Performing absolute dosimetry resulted in a correlation coefficient of R=0.9916(pâ€value≈10−7) between standard interstitial applicator and the one proposed in this article. This study not only introduces a novel applicator with acceptable attenuation but also proves the response independency of the GAFCHROMIC EBT films to energy. By applying the dose response of the applicator in the treatment planning software, it can be used as a new intraluminal / interstitial applicator. PACS number: 87.53.Bn, 87.53.Jw, 29.40.Cs

  9. Interstitial high-dose rate brachytherapy as boost for anal canal cancer

    International Nuclear Information System (INIS)

    Falk, Alexander Tuan; Claren, Audrey; Benezery, Karen; François, Eric; Gautier, Mathieu; Gerard, Jean-Pierre; Hannoun-Levi, Jean-Michel

    2014-01-01

    To assess clinical outcomes of patients treated with a high-dose rate brachytherapy boost for anal canal cancer (ACC). From August 2005 to February 2013, 28 patients presenting an ACC treated by split-course external beam radiotherapy (EBRT) and HDR brachytherapy with or without chemotherapy in a French regional cancer center in Nice were retrospectively analyzed. Median age was 60.6 years [34 – 83], 25 patients presented a squamous cell carcinoma and 3 an adenocarcinoma; 21 received chemotherapy. Median dose of EBRT was 45 Gy [43.2 – 52]. Median dose of HDR brachytherapy was 12 Gy [10 - 15] with a median duration of 2 days. Median overall treatment time was 63 days and median delay between EBRT and brachytherapy was 20 days. Two-year local relapse free, metastatic free, disease free and overall survivals were 83%, 81.9%, 71.8% and 87.7% respectively. Acute toxicities were frequent but not severe with mostly grade 1 toxicities: 37% of genito-urinary, 40.7% of gastro-intestinal and 3.7% of cutaneous toxicities. Late toxicities were mainly G1 (43.1%) and G2 (22%). Two-year colostomy-free survival was 75.1%, one patient had a definitive sphincter amputation. High-dose rate brachytherapy for anal canal carcinoma as boost represents a feasible technique compared to low or pulsed-dose rate brachytherapy. This technique remains an excellent approach to precisely boost the tumor in reducing the overall treatment time

  10. Balloon-based adjuvant radiotherapy in breast cancer: comparison between 99mTc and HDR 192Ir

    Directory of Open Access Journals (Sweden)

    Tarcísio Passos Ribeiro de Campos

    2016-04-01

    Full Text Available Abstract Objective: To perform a comparative dosimetric analysis, based on computer simulations, of temporary balloon implants with 99mTc and balloon brachytherapy with high-dose-rate (HDR 192Ir, as boosts to radiotherapy. We hypothesized that the two techniques would produce equivalent doses under pre-established conditions of activity and exposure time. Materials and Methods: Simulations of implants with 99mTc-filled and HDR 192Ir-filled balloons were performed with the Siscodes/MCNP5, modeling in voxels a magnetic resonance imaging set related to a young female. Spatial dose rate distributions were determined. In the dosimetric analysis of the protocols, the exposure time and the level of activity required were specified. Results: The 99mTc balloon presented a weighted dose rate in the tumor bed of 0.428 cGy.h-1.mCi-1 and 0.190 cGyh-1.mCi-1 at the balloon surface and at 8-10 mm from the surface, respectively, compared with 0.499 and 0.150 cGyh-1.mCi-1, respectively, for the HDR 192Ir balloon. An exposure time of 24 hours was required for the 99mTc balloon to produce a boost of 10.14 Gy with 1.0 Ci, whereas only 24 minutes with 10.0 Ci segments were required for the HDR 192Ir balloon to produce a boost of 5.14 Gy at the same reference point, or 10.28 Gy in two 24-minutes fractions. Conclusion: Temporary 99mTc balloon implantation is an attractive option for adjuvant radiotherapy in breast cancer, because of its availability, economic viability, and similar dosimetry in comparison with the use of HDR 192Ir balloon implantation, which is the current standard in clinical practice.

  11. Curative high dose rate vaginal apex brachytherapy in stage I papillary serous carcinoma of the endometrium

    International Nuclear Information System (INIS)

    Turner, B.C.; Kacinski, B.M.; Gumbs, A.; Peschel, R.E.; Haffty, B.G.; Wilson, L.D.

    1996-01-01

    Introduction: Uterine papillary serous carcinoma (UPSC) is a morphologically distinct variant of endometrial carcinoma that is associated with a poor prognosis, high recurrence rate, clinical understaging, and poor response to salvage treatment. We describe the presentation, local and distant control, survival, salvage rate, and complications for patients undergoing whole abdominal radiation therapy (WART), low dose rate (LDR) intracavitary brachytherapy, or high dose rate (HDR) vaginal brachytherapy in patients with stage I UPSC. Methods: Between 1976 and 1994 more than 1700 patients with endometrial carcinoma were treated with radiation therapy, 30 patients with stage I UPSC (1.8%) were treated with radiation before or following TAH/BSO. All patients underwent either preoperative Simon's packing or tandem and plaque which delivered 30-40 Gy to the serosa, WART, or HDR Ir-192 vaginal apex brachytherapy to a total dose of 21 Gy in 3 fractions at 0.5 cm from the vaginal mucosa. A total of 14 patients received HDR vaginal brachytherapy and (5(14)) patients received systemic chemotherapy. All patients presented with vaginal bleeding at a median age of 67 years (range 34-88). The group of 30 patients underwent TAH/BSO, 17 patients were completely staged pathologically (pelvic and para-aortic lymph nodes, omentectomy, and pelvic washings), and 2 patients underwent omental biopsy and pelvic washings only. All specimens revealed UPSC, nuclear grade 3, and lymphovascular invasion (23%). The pathologic stage was IA: 23% (7), IB: 67% (20), and IC: 10% (3). The median follow-up for all patients was 49 months (range 13-187 months). For the patients receiving postoperative HDR vaginal brachytherapy the median time from surgery to radiation was 42 days (range 29-91). Results: The 5-year actuarial disease free survival for Figo stage I UPSC patients treated with postoperative HDR vaginal brachytherapy and systemic chemotherapy was 100% compared to 74% for stage I UPSC patient

  12. Characterization of HDR Ir-192 source for 3D planning system

    International Nuclear Information System (INIS)

    Fonseca, Gabriel P.; Yoriyaz, Helio; Antunes, Paula C.G.; Siqueira, Paulo T.D.; Rubo, Rodrigo; Ferreira, Louise A.

    2011-01-01

    Brachytherapy treatment involves surgical or cavitary insertion of radioactive sources for diseases treatments, such as: lung, gynecologic or prostate cancer. This technique has great ability to administer high doses to the tumor, with adjacent normal tissue preservation equal or better than external beam radiation therapy. Several innovations have been incorporated in this treatment technique, such as, 3D treatment planning system and computer guided sources. In detriment to scientific advances there are no protocols that relate dose with tumor volume, organs or A point, established by ICRU38 and used to prescribe dose in treatment planning system. Several international studies, like as EMBRACE, the multicentre international study, has been trying to correlate the dose volume using 3D planning systems and medical images, as those obtained by CT or MRI, to establish treatment protocols. With the objective of analyzing the 3D dose distribution, a micro Selectron-HDR remote afterloading device for high dose-rate (HDR) was characterized in the present work. Through the data provided by the manufacturer the source was simulated, using the MCNP5 code to calculate American Association of Physicists in Medicine Task Group No. 43 report (AAPM TG43) specified parameters. The simulations have shown great agreement when compared to the ONCENTRA planning system results and those provided by literature. The micro Selectron-HDR remote afterloading device will be utilized to simulate 3D dose distribution through CT images processed by an auxiliary software which process DICOM images. (author)

  13. Characterization of HDR Ir-192 source for 3D planning system

    Energy Technology Data Exchange (ETDEWEB)

    Fonseca, Gabriel P.; Yoriyaz, Helio; Antunes, Paula C.G.; Siqueira, Paulo T.D., E-mail: gabriel.fonseca@usp.b, E-mail: hyoriyaz@ipen.b, E-mail: ptsiquei@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Rubo, Rodrigo [Universidade de Sao Paulo (HC/FMUSP), Sao Paulo, SP (Brazil). Hospital das Clinicas. Servico de Radioterapia; Minamisawa, Renato A., E-mail: renato.minamisawa@psi.c [Paul Scherrer Institut (PSI), Villigen (Switzerland); Ferreira, Louise A. [Universidade Estadual de Maringa (UEM), PR (Brazil). Fac. de Medicina

    2011-07-01

    Brachytherapy treatment involves surgical or cavitary insertion of radioactive sources for diseases treatments, such as: lung, gynecologic or prostate cancer. This technique has great ability to administer high doses to the tumor, with adjacent normal tissue preservation equal or better than external beam radiation therapy. Several innovations have been incorporated in this treatment technique, such as, 3D treatment planning system and computer guided sources. In detriment to scientific advances there are no protocols that relate dose with tumor volume, organs or A point, established by ICRU38 and used to prescribe dose in treatment planning system. Several international studies, like as EMBRACE, the multicentre international study, has been trying to correlate the dose volume using 3D planning systems and medical images, as those obtained by CT or MRI, to establish treatment protocols. With the objective of analyzing the 3D dose distribution, a micro Selectron-HDR remote afterloading device for high dose-rate (HDR) was characterized in the present work. Through the data provided by the manufacturer the source was simulated, using the MCNP5 code to calculate American Association of Physicists in Medicine Task Group No. 43 report (AAPM TG43) specified parameters. The simulations have shown great agreement when compared to the ONCENTRA planning system results and those provided by literature. The micro Selectron-HDR remote afterloading device will be utilized to simulate 3D dose distribution through CT images processed by an auxiliary software which process DICOM images. (author)

  14. Radioablation of adrenal gland malignomas with interstitial high-dose-rate brachytherapy. Efficacy and outcome

    Energy Technology Data Exchange (ETDEWEB)

    Mohnike, K. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Radiologie und Nuklearmedizin, Magdeburg (Germany); DTZ am Frankfurter Tor, Berlin (Germany); Neumann, K.; Seidensticker, M.; Seidensticker, R.; Pech, M.; Streitparth, T.; Ricke, J. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Radiologie und Nuklearmedizin, Magdeburg (Germany); Hass, P.; Gademann, G. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Strahlentherapie, Magdeburg (Germany); Klose, S. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Nieren- und Hochdruckkrankheiten, Diabetologie und Endokrinologie, Magdeburg (Germany); Garlipp, B.; Benckert, C. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Allgemein-, Viszeral- und Gefaesschirurgie, Magdeburg (Germany); Wendler, J.J.; Liehr, U.B.; Schostak, M. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Urologie und Kinderurologie, Magdeburg (Germany); Goeppner, D. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Dermatologie, Magdeburg (Germany)

    2017-08-15

    To assess the efficacy, safety, and outcome of image-guided high-dose-rate (HDR) brachytherapy in patients with adrenal gland metastases (AGM). From January 2007 to April 2014, 37 patients (7 female, 30 male; mean age 66.8 years, range 41.5-82.5 years) with AGM from different primary tumors were treated with CT-guided HDR interstitial brachytherapy (iBT). Primary endpoint was local tumor control (LTC). Secondary endpoints were time to untreatable progression (TTUP), time to progression (TTP), overall survival (OS), and safety. In a secondary analysis, risk factors with an influence on survival were identified. The median biological equivalent dose (BED) was 37.4 Gy. Mean LTC after 12 months was 88%; after 24 months this was 74%. According to CTCAE criteria, one grade 3 adverse event occurred. Median OS after first diagnosis of AGM was 18.3 months. Median OS, TTUP, and TTP after iBT treatment were 11.4, 6.6, and 3.5 months, respectively. Uni- and multivariate Cox regression analyses revealed significant influences of synchronous disease, tumor diameter, and the total number of lesions on OS or TTUP or both. Image-guided HDR-iBT is safe and effective. Treatment- and primary tumor-independent features influenced survival of patients with AGM after HDR-iBR treatment. (orig.) [German] Beurteilung der Effektivitaet, Sicherheit und Ergebnisse nach bildgefuehrter High-dose-rate-(HDR-)Brachytherapie bei Patienten mit Nebennierenmetastasen. Von Januar 2007 bis April 2014 wurden 37 Patienten (7 weiblich, 30 maennlich; mittleres Alter 66,8 Jahre, Spanne 41,5-82,5 Jahre) mit Nebennierenmetastasen verschiedener Primarien mit CT-gesteuerter interstitieller HDR-Brachytherapie (iBT) behandelt. Der primaere Endpunkt war die lokale Tumorkontrolle (LTC). Sekundaere Endpunkte umfassten die Zeit bis zum nicht mehr behandelbaren Progress (TTUP), die Zeit bis zum Progress (TTP), das Gesamtueberleben (OS) und die Sicherheit der Methode. In einer sekundaeren Analyse wurden Risikofaktoren

  15. High dose rate brachytherapy for carcinoma of the cervix: risk factors for late rectal complications

    International Nuclear Information System (INIS)

    Uno, Takashi; Itami, Jun; Aruga, Moriyo; Kotaka, Kikuo; Fujimoto, Hajime; Minoura, Shigeki

    1996-01-01

    Purpose/Objective: To determine the incidence of late rectal complications in patients treated with high dose rate brachytherapy for FIGO stage IIB, IIIB carcinoma of the uterine cervix, and to evaluate the treatment factors associated with an increased probability of treatment complications. Materials and Methods: Records of 100 patients with FIGO IIB or IIIB cervical carcinoma treated with definitive irradiation using high dose rate intracavitary brachytherapy (HDR-ICR) between 1977 and 1994 were retrospectively reviewed. For each HDR-ICR session, 6 Gy isodose volume was reconstructed three dimensionally and the following three parameters were determined to represent this isodose volume, length (L); maximum longitudinal distance of 6 Gy isodose area in an oblique frontal plane containing the intrauterine applicator, width (W); maximum width of 6 Gy isodose area in the same plane, height (H); maximum dimension of 6 Gy isodose area perpendicular to the intrauterine applicator determined in the oblique sagittal plane. Point P/Q (2 cm ventral/dorsal from the proximal retention point of the intrauterine source) and point R/S (2 cm ventral/dorsal from the midpoint of the ovoid sources) were also defined retrospectively and HDR-ICR dose at these points were calculated. Statistical analyses were performed to determine the treatment factors predictive of late rectal complications. Results: The 5-year cumulative cause-specific disease-free survival rate was 50% for all, 74% for Stage IIB, and 38% for Stage IIIB, with a significant difference between two FIGO Stages (p=0.0004). Of patients treated for both stages, 30% and 36% had experienced moderate to severe (Grade 2-4) complications at 3 and 5 years, respectively. Average H value (p=0.013) and cumulative point S dose by HDR-ICR (p=0.020) were significantly correlated with the incidence of late rectal complications (Student's t-test), whereas these factors did not significantly affect the probability of pelvic control. No

  16. Effectiveness of two different HDR brachytherapy regimens with the same BED value in cervical cancer

    Directory of Open Access Journals (Sweden)

    Rajesh Vashistha

    2010-07-01

    Full Text Available Purpose: To analyze the effectiveness of biologically effective dose (BED in two different regimens of HDR brachytherapy keeping the same total BED to point A and to compare the relationship of overall treatment time in terms of local control and bladder and rectal complications.Material and methods: The study included two groups comprising a total of 90 cervical cancer patients who underwent external beam radiotherapy (EBRT followed by HDR intracavitary brachytherapy (ICBT. EBRT treatment was delivered by a Co-60 teletherapy unit to a prescribed dose of 45 Gy with 1.8 Gy per fraction in 25 fractions over a period of five weeks. Parallel opposed anterior–posterior (AP/PA fields with no central shielding were used, followed by the HDR ICBT dose, to point A, of either two fractions of 9.5 Gy with a gap of 10 days, or three fractions of 7.5 Gy with a gap of 7 days between the fractions. Gemcitabine (dose of 150 mg/m2 was given weekly to all the patients as a radiosensitizer. The calculate BED3 to point A was almost the same in both groups to keep the same late complication rates. The doses, and BED10 and BED3, were calculated at different bladder and rectal point as well as at the lymphatictrapezoid points. During and after treatment patients were evaluated for local control and complications for 24 months.Results and Conclusions: Doses and BEDs at different bladder, rectal and lymphatic trapezoid points, local control, and complications in both HDR ICBT groups did not have statistically significant differences (p > 0.05. Both HDR ICBT schedules are well tolerable and equally effective.

  17. Updated results of high-dose rate brachytherapy and external beam radiotherapy for locally and locally advanced prostate cancer using the RTOG-ASTRO phoenix definition

    Directory of Open Access Journals (Sweden)

    Antonio C. Pellizzon

    2008-06-01

    Full Text Available PURPOSE: To evaluate the prognostic factors for patients with local or locally advanced prostate cancer treated with external beam radiotherapy (RT and high dose rate brachytherapy (HDR according to the RTOG-ASTRO Phoenix Consensus Conference. MATERIALS AND METHODS: The charts of 209 patients treated between 1997 and 2005 with localized RT and HDR as a boost at the Department of Radiation Oncology, AC Camargo Hospital, Sao Paulo, Brazil were reviewed. Clinical and treatment parameters i.e.: patient's age, Gleason score, clinical stage, initial PSA (iPSA, risk group (RG for biochemical failure, doses of RT and HDR were evaluated. Median age and median follow-up time were 68 and 5.3 years, respectively. Median RT and HDR doses were 45 Gy and 20 Gy. RESULTS: Disease specific survival (DSS at 3.3 year was 94.2%. Regarding RG, for the LR (low risk, IR (intermediate risk and HR (high risk, the DSS rates at 3.3 years were 91.5%, 90.2% and 88.5%, respectively. On univariate analysis prognostic factors related to DSS were RG (p = 0.040, Gleason score ≤ 6 ng/mL (p = 0.002, total dose of HDR ≥ 20 Gy (p < 0.001 On multivariate analysis the only statistical significant predictive factor for biochemical control (bNED was the RG, p < 0.001 (CI - 1.147-3.561. CONCLUSIONS: Although the radiation dose administered to the prostate is an important factor related to bNED, this could not be established with statistical significance in this group of patients. To date , in our own experience, HDR associated to RT could be considered a successful approach in the treatment of prostate cancer.

  18. Source position verification and dosimetry in HDR brachytherapy using an EPID

    International Nuclear Information System (INIS)

    Smith, R. L.; Taylor, M. L.; McDermott, L. N.; Franich, R. D.; Haworth, A.; Millar, J. L.

    2013-01-01

    Purpose: Accurate treatment delivery in high dose rate (HDR) brachytherapy requires correct source dwell positions and dwell times to be administered relative to each other and to the surrounding anatomy. Treatment delivery inaccuracies predominantly occur for two reasons: (i) anatomical movement or (ii) as a result of human errors that are usually related to incorrect implementation of the planned treatment. Electronic portal imaging devices (EPIDs) were originally developed for patient position verification in external beam radiotherapy and their application has been extended to provide dosimetric information. The authors have characterized the response of an EPID for use with an 192 Ir brachytherapy source to demonstrate its use as a verification device, providing both source position and dosimetric information.Methods: Characterization of the EPID response using an 192 Ir brachytherapy source included investigations of reproducibility, linearity with dose rate, photon energy dependence, and charge build-up effects associated with exposure time and image acquisition time. Source position resolution in three dimensions was determined. To illustrate treatment verification, a simple treatment plan was delivered to a phantom and the measured EPID dose distribution compared with the planned dose.Results: The mean absolute source position error in the plane parallel to the EPID, for dwells measured at 50, 100, and 150 mm source to detector distances (SDD), was determined to be 0.26 mm. The resolution of the z coordinate (perpendicular distance from detector plane) is SDD dependent with 95% confidence intervals of ±0.1, ±0.5, and ±2.0 mm at SDDs of 50, 100, and 150 mm, respectively. The response of the EPID is highly linear to dose rate. The EPID exhibits an over-response to low energy incident photons and this nonlinearity is incorporated into the dose calibration procedure. A distance (spectral) dependent dose rate calibration procedure has been developed. The

  19. Biological effect of pulsed dose rate brachytherapy with stepping sources if short half-times of repair are present in tissues

    International Nuclear Information System (INIS)

    Fowler, Jack F.; Limbergen, Erik F.M. van

    1997-01-01

    Purpose: To explore the possible increase of radiation effect in tissues irradiated by pulsed brachytherapy (PDR) for local tissue dose rates between those 'averaged over the whole pulse' and the instantaneous high dose rates close to the dwell positions. Increased effect is more likely for tissues with short half-times of repair of the order of a few minutes, similar to pulse durations. Methods and Materials: Calculations were done assuming the linear quadratic formula for radiation damage, in which only the dose-squared term is subject to exponential repair. The situation with two components of T (1(2)) is addressed. A constant overall time of 140 h and a constant total dose of 70 Gy were assumed throughout, the continuous low dose rate of 0.5 Gy/h (CLDR) providing the unitary standard effects for each PDR condition. Effects of dose rates ranging from 4 Gy/h to 120 Gy/h (HDR at 2 Gy/min) were studied, covering the gap in an earlier publication. Four schedules were examined: doses per pulse of 0.5, 1, 1.5, and 2 Gy given at repetition frequencies of 1, 2, 3, and 4 h, respectively, each with a range of assumed half-times of repair of 4 min to 1.5 h. Results are presented for late-responding tissues, the differences from CLDR being two or three times greater than for early-responding tissues and most tumors. Results: Curves are presented relating the ratio of increased biological effect (proportional to log cell kill) calculated for PDR relative to CLDR. Ratios as high as 1.5 can be found for large doses per pulse (2 Gy) if the half-time of repair in tissues is as short as a few minutes. The major influences on effect are dose per pulse, half-time of repair in tissue, and--when T (1(2)) is short--the instantaneous dose rate. Maximum ratios of PDR/CLDR occur when the dose rate is such that pulse duration is approximately equal to T (1(2)) . As dose rate in the pulse is increased, a plateau of effect is reached, for most T (1(2)) s, above 10 to 20 Gy/h, which is

  20. In phantom calibration of a high dose rate remote afterloading device

    International Nuclear Information System (INIS)

    Alfonso, R.; Tolede, P.; Pich, V.

    1995-01-01

    The high dose-rate (HDR) brachytherapy in Cuba is based on soviet made devices type AGAT-V. In order to calibration one of these for clinical use a method based of the different measurement of absorbed dose at the reference point B in a paraffin phantom was developed. The results of the calibration are shown. From these results an analysis was made of the effective doses to prescription point a considering the Lineal-Quadratic model. The clinical results by using the AGAT-V device are displayed in a comparative way

  1. Virtual HDR CyberKnife SBRT for Localized Prostatic Carcinoma: 5-year Disease-free Survival and Toxicity Observations

    Directory of Open Access Journals (Sweden)

    Donald Blake Fuller

    2014-11-01

    Full Text Available PURPOSEProstate stereotactic body radiotherapy (SBRT may substantially recapitulate the dose distribution of high-dose-rate (HDR brachytherapy, representing an externally delivered Virtual HDR treatment method. Herein we present 5-year outcomes from a cohort of consecutively treated Virtual HDR SBRT prostate cancer patients.METHODSSeventy-nine patients were treated from 2006 - 2009, 40 low-risk and 39 intermediate-risk, under IRB-approved clinical trial, to 38 Gy in 4 fractions. The planning target volume (PTV included prostate plus a 2-mm volume expansion in all directions, with selective use of a 5-mm prostate-to-PTV expansion and proximal seminal vesicle coverage in intermediate-risk patients, to better cover potential extraprostatic disease; rectal PTV margin reduced to zero in all cases. The prescription dose covered > 95% of the PTV (V100 >= 95%, with a minimum 150% PTV dose escalation to create HDR-like PTV dose distribution.RESULTSMedian pre-SBRT PSA level of 5.6 ng/mL decreased to 0.05 ng/mL 5 years out and 0.02 ng/mL 6 years out. At least one PSA bounce was seen in 55 patients (70% but only 3 of them subsequently relapsed, Biochemical-relapse-free survival was 100% and 92% for low-risk and intermediate-risk patients, respectively, by ASTRO definition (98% and 92% by Phoenix definition. Local relapse did not occur, distant metastasis-free survival was 100% and 95% by risk-group, and disease-specific survival was 100%. Acute and late grade 2 GU toxicity incidence was 10% and 9%, respectively; with 6% late grade 3 GU toxicity. Acute urinary retention did not occur. Acute and late grade 2 GI toxicity was 0% and 1%, respectively, with no grade 3 or higher toxicity. Of patients potent pre-SBRT, 65% remained so at 5 years.CONCLUSIONSVirtual HDR prostate SBRT creates a very low PSA nadir, a high rate of 5-year disease-free survival and an acceptable toxicity incidence, with results closely resembling those reported post-HDR brachytherapy.

  2. High dose rate brachytherapy for medically inoperable stage I endometrial cancer

    Energy Technology Data Exchange (ETDEWEB)

    Petereit, Daniel G; Sarkaria, Jann N; Schink, Julian; Springman, Scott R; Kinsella, Timothy J; Buchler, Dolores A

    1995-07-01

    Purpose/Objective: To determine the efficacy of high dose rate (HDR) brachytherapy in patients with medically inoperable endometrial cancer clinically confined to the corpus. Materials and Methods: Forty-two patients with endometrial cancer and an intact uterus have been treated since 1989 with HDR brachytherapy. Twenty-six patients with medically inoperable Stage I disease were treated with radiation alone and form the basis of this study. Obesity was assessed using the body mass index (BMI kg/m{sup 2}) scale. Patients with a BMI above 28 were considered obese and those above 35 morbidly obese, per standard anesthesia guidelines. Brachytherapy was delivered in 5 HDR insertions, 1 week apart, without any external beam radiation. The following doses were delivered per insertion: 5.7 Gy to point S, 7.0 Gy to point W, 8.2 Gy to the vaginal surface and 9.2 Gy to point M. Point M represents the conventional point A dose, while points S and W are myometrial points. A single tandem with either ovoids or cylinders was placed, unless the uterine cavity would accommodate 2 tandems. All treatments were outpatient using intravenous fentanyl and midazolam for sedation. Pelvic ultrasound was commonly used at the time of brachytherapy to verify tandem placement. Three year clinical endpoints were calculated using the Kaplan Meier method. Results: The median follow-up for the study cohort was 21 months with follow-up greater than 36 months in 11 patients. Seventeen of the 26 patients were inoperable due to morbid obesity (median weight and BMI; 316 lbs and 55 kg/m{sup 2}, respectively); the other patients had poor cardiopulmonary reserve {+-} obesity. The median age, KPS (Karnofsky Performance Status), weight, ASA (American Society of Anesthesiologists' Physical Class System) and BMI were 63 yrs, 80%, 285 lbs, 3 and 49 kg/m{sup 2}, respectively. Two patients with an ASA of 3 and 4 died from acute cardio-pulmonary events within 30 days of the last insertion, emphasizing the need

  3. High dose rate brachytherapy for medically inoperable stage I endometrial cancer

    International Nuclear Information System (INIS)

    Petereit, Daniel G.; Sarkaria, Jann N.; Schink, Julian; Springman, Scott R.; Kinsella, Timothy J.; Buchler, Dolores A.

    1995-01-01

    Purpose/Objective: To determine the efficacy of high dose rate (HDR) brachytherapy in patients with medically inoperable endometrial cancer clinically confined to the corpus. Materials and Methods: Forty-two patients with endometrial cancer and an intact uterus have been treated since 1989 with HDR brachytherapy. Twenty-six patients with medically inoperable Stage I disease were treated with radiation alone and form the basis of this study. Obesity was assessed using the body mass index (BMI kg/m 2 ) scale. Patients with a BMI above 28 were considered obese and those above 35 morbidly obese, per standard anesthesia guidelines. Brachytherapy was delivered in 5 HDR insertions, 1 week apart, without any external beam radiation. The following doses were delivered per insertion: 5.7 Gy to point S, 7.0 Gy to point W, 8.2 Gy to the vaginal surface and 9.2 Gy to point M. Point M represents the conventional point A dose, while points S and W are myometrial points. A single tandem with either ovoids or cylinders was placed, unless the uterine cavity would accommodate 2 tandems. All treatments were outpatient using intravenous fentanyl and midazolam for sedation. Pelvic ultrasound was commonly used at the time of brachytherapy to verify tandem placement. Three year clinical endpoints were calculated using the Kaplan Meier method. Results: The median follow-up for the study cohort was 21 months with follow-up greater than 36 months in 11 patients. Seventeen of the 26 patients were inoperable due to morbid obesity (median weight and BMI; 316 lbs and 55 kg/m 2 , respectively); the other patients had poor cardiopulmonary reserve ± obesity. The median age, KPS (Karnofsky Performance Status), weight, ASA (American Society of Anesthesiologists' Physical Class System) and BMI were 63 yrs, 80%, 285 lbs, 3 and 49 kg/m 2 , respectively. Two patients with an ASA of 3 and 4 died from acute cardio-pulmonary events within 30 days of the last insertion, emphasizing the need for accurate pre

  4. Impact of 'optimized' treatment planning for tandem and ring, and tandem and ovoids, using high dose rate brachytherapy for cervical cancer

    International Nuclear Information System (INIS)

    Noyes, William R.; Peters, Nancy E.; Thomadsen, Bruce R.; Fowler, Jack F.; Buchler, Dolores A.; Stitt, Judith A.; Kinsella, Timothy J.

    1995-01-01

    Purpose: Different treatment techniques are used in high dose rate (HDR) remote afterloading intracavitary brachytherapy for uterine cervical cancer. We have investigated the differences between 'optimized' and 'nonoptimized' therapy using both a tandem and ring (T/R) applicator, and a tandem and ovoids (T/O), applicator. Methods and Materials: HDR afterloading brachytherapy using the Madison System for Stage IB cervical cancer was simulated for 10 different patients using both a T/R applicator and a T/O applicator. A treatment course consists of external beam irradiation and five insertions of HDR afterloading brachytherapy. Full dosimetry calculations were performed at the initial insertion for both applicators and used as a reference for the following four insertions of the appropriate applicator. Forty dosimetry calculations were performed to determine the dose delivered to Point M (similar to Point A), Point E (obturator lymph nodes), vaginal surface, bladder, and rectum. 'Optimized' doses were specified to Point M and to the vaginal surface. 'Nonoptimized' doses were specified to Point M only. Using the linear-quadratic equation, calculations have been performed to convert the delivered dose using HDR to the biologically equivalent doses at the conventional low dose rate (LDR) at 0.60 Gy/h. Results: Major differences between 'optimized' and 'nonoptimized' LDR equivalent doses were found at the vaginal surface, bladder, and rectum. Overdoses at the vaginal surface, bladder, and rectum were calculated to be 208%, nil, and 42%, respectively, for the T/R applicator with 'nonoptimization'. However, for the T/O applicator, the overdoses were smaller, being nil, 32%, and 27%, respectively, with 'nonoptimization'. Conclusion: Doses given in high dose rate intracavitary brachytherapy border on tissue tolerance. 'Optimization' of either applicator decreases the risk of a dose that may have potential for complications. Optimization of a tandem and ovoids best ensures

  5. High-dose-rate-intracavitary brachytherapy applications and the difference in the bladder and rectum doses: A study from rural centre of Maharashatra, India

    Directory of Open Access Journals (Sweden)

    Jain Vandana

    2007-01-01

    Full Text Available Aim : To report the difference in the bladder and rectum doses with different applications by the radiotherapists in the same patient of the carcinoma of the uterine cervix treated by multiple fractions of high-dose-rate (HDR intracavitary brachytherapy (ICBT. Materials and Methods : Between January 2003 to December 2004, a total of 60 cases of the carcinoma uterine cervix were selected randomly for the retrospective analyses. All 60 cases were grouped in six groups according to the treating radiotherapist who did the HDR-ICBT application. Three radiotherapists were considered for this study, named A, B and C. Ten cases for each radiotherapist in whom all three applications were done by the same radiotherapist. And 10 cases for each radiotherapist with shared applications in the same patient (A+B, A+C and B+C. The bladder and rectal doses were calculated in reference to point "A" dose and were limited to 80% of prescribed point "A" dose, as per ICRU-38 recommendations. Received dose grouped in three groups- less then 80% (< 80%, 80-100% and above 100% (>100%. A total of 180 applications for 60 patients were calculated for the above analyses. Results : There is a lot of difference in the bladder and rectal doses with the application by the different radiotherapists, even in the same patient with multiple fractions of HDR-ICBT. Applications by ′A′ radiotherapist were within the limits in the self as well as in the shared groups more number of times, by ′B′ radiotherapist was more times exceeding the limit and by ′C′ radiotherapist doses were in between the A and B. Discussion and Conclusion : For the rectal and bladder doses most important factors are patient′s age, disease stage, duration between EBRT and HDR-ICRT and patient anatomy, but these differences can be minimized to some extent by careful application, proper packing and proper fixation.

  6. High-dose-rate brachytherapy as salvage modality for locally recurrent prostate cancer after definitive radiotherapy. A systematic review

    Energy Technology Data Exchange (ETDEWEB)

    Chatzikonstantinou, Georgios; Zamboglou, Nikolaos; Roedel, Claus; Tselis, Nikolaos [J.W. Goethe University of Frankfurt, Department of Radiotherapy and Oncology, Frankfurt am Main (Germany); Zoga, Eleni [Sana Klinikum Offenbach, Department of Radiotherapy and Oncology, Offenbach am Main (Germany); Strouthos, Iosif [Medical Center - University of Freiburg, Department of Radiotherapy and Oncology, University of Freiburg, Freiburg (Germany); Butt, Saeed Ahmed [Sana Klinikum Offenbach, Department of Medical Physics and Engineering, Offenbach am Main (Germany)

    2017-09-15

    To review the current status of interstitial high-dose-rate brachytherapy as a salvage modality (sHDR BRT) for locally recurrent prostate cancer after definitive radiotherapy (RT). A literature search was performed in PubMed using ''high-dose-rate, brachytherapy, prostate cancer, salvage'' as search terms. In all, 51 search results published between 2000 and 2016 were identified. Data tables were generated and summary descriptions created. The main outcome parameters used were biochemical control (BC) and toxicity scores. Eleven publications reported clinical outcome and toxicity with follow-up ranging from 4-191 months. A variety of dose and fractionation schedules were described, including 19.0 Gy in 2 fractions up to 42.0 Gy in 6 fractions. The 5-year BC ranged from 18-77%. Late grade 3 genitourinary and gastrointestinal toxicity was 0-32% and 0-5.1%, respectively. sHDR BRT appears as safe and effective salvage modality for the reirradiation of locally recurrent prostate cancer after definitive RT. (orig.) [German] Zusammenfassende Darstellung relevanter Literatur zur interstitiellen High-Dose-Rate-Brachytherapie als Salvage-Modalitaet (sHDR-BRT) bei der Behandlung des lokal rezidivierten Prostatakarzinoms nach vorausgegangener definitiver Radiotherapie (RT). In der PubMed-Datenbank wurde eine Literaturrecherche mit den Suchbegriffen ''high-dose-rate, brachytherapy, prostate cancer, salvage'' durchgefuehrt. Zwischen den Jahren 2000 und 2016 wurden 51 Publikationen identifiziert. Die biochemische Kontrolle (BC) sowie das assoziierte Toxizitaetsprofil waren onkologische Hauptpunkte in der Analyse der beruecksichtigten Literatur. Von onkologischen Ergebnissen und Toxizitaeten berichteten 11 Publikationen bei einer medianen Nachbeobachtungszeit von 4-191 Monaten. Eine Variabilitaet von Dosis- und Fraktionierungsregimen wurde beschrieben mit totalen physikalischen Dosen von 19,0 Gy in 2 Fraktionen bis zu 42,0 Gy in 6 Fraktionen

  7. Balloon-based adjuvant radiotherapy in breast cancer: comparison between {sup 99m}Tc and HDR {sup 192}Ir

    Energy Technology Data Exchange (ETDEWEB)

    Campos, Tarcisio Passos Ribeiro de; Lima, Carla Flavia de; Cuperschmid, Ethel Mizrahy, E-mail: tprcampos@pq.cnpq.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil)

    2016-03-15

    Objective: To perform a comparative dosimetric analysis, based on computer simulations, of temporary balloon implants with {sup 99m}Tc and balloon brachytherapy with high-dose-rate (HDR) {sup 192}Ir, as boosts to radiotherapy. We hypothesized that the two techniques would produce equivalent doses under pre-established conditions of activity and exposure time. Materials and methods: simulations of implants with {sup 99m}Tc-filled and HDR {sup 192}Ir-filled balloons were performed with the Siscodes/MCNP5, modeling in voxels a magnetic resonance imaging set related to a young female. Spatial dose rate distributions were determined. In the dosimetric analysis of the protocols, the exposure time and the level of activity required were specified. Results: the {sup 99m}Tc balloon presented a weighted dose rate in the tumor bed of 0.428 cGy.h{sup -1}.mCi{sup -1} and 0.190 cGyh{sup -1} at the balloon surface and at 8-10 mm from the surface, respectively, compared with 0.499 and 0.150 cGyh{sup -1}.mCi{sup -1}, respectively, for the HDR {sup 192}Ir balloon. An exposure time of 24 hours was required for the {sup 99m}Tc balloon to produce a boost of 10.14 Gy with 1.0 Ci, whereas only 24 minutes with 10.0 Ci segments were required for the HDR {sup 192}Ir balloon to produce a boost of 5.14 Gy at the same reference point, or 10.28 Gy in two 24-minutes fractions. Conclusion: temporary {sup 99m}Tc balloon implantation is an attractive option for adjuvant radiotherapy in breast cancer, because of its availability, economic viability, and similar dosimetry in comparison with the use of HDR {sup 192}Ir balloon implantation, which is the current standard in clinical practice. (author)

  8. HDR brachytherapy for superficial non-melanoma skin cancers

    International Nuclear Information System (INIS)

    Gauden, Ruth; Pracy, Martin; Avery, Anne-Marie; Hodgetts, Ian; Gauden, Stan

    2013-01-01

    Our initial experience using recommended high dose per fraction skin brachytherapy (BT) treatment schedules, resulted in poor cosmesis. This study aimed to assess in a prospective group of patients the use of Leipzig surface applicators for High Dose Rate (HDR) brachytherapy, for the treatment of small non-melanoma skin cancers (NMSC) using a protracted treatment schedule. Treatment was delivered by HDR brachytherapy with Leipzig applicators. 36Gy, prescribed to between 3 to 4mm, was given in daily 3Gy fractions. Acute skin toxicity was evaluated weekly during irradiation using the Radiation Therapy Oncology Group criteria. Local response, late skin effects and cosmetic results were monitored at periodic intervals after treatment completion. From March 2002, 200 patients with 236 lesions were treated. Median follow-up was 66 months (range 25–121 months). A total of 162 lesions were macroscopic, while in 74 cases, BT was given after resection because of positive microscopic margins. There were 121 lesions that were basal cell carcinomas, and 115 were squamous cell carcinomas. Lesions were located on the head and neck (198), the extremities (26) and trunk (12). Local control was 232/236 (98%). Four patients required further surgery to treat recurrence. Grade 1 acute skin toxicity was detected in 168 treated lesions (71%) and grade 2 in 81 (34%). Cosmesis was good or excellent in 208 cases (88%). Late skin hypopigmentation changes were observed in 13 cases (5.5%). Delivering 36Gy over 2 weeks to superficial NMSC using HDR brachytherapy is well tolerated and provides a high local control rate without significant toxicity.

  9. Adjuvant high-dose-rate brachytherapy after external beam radiotherapy in nasopharyngeal carcinoma

    International Nuclear Information System (INIS)

    Oezyar, Enis; Yildz, Ferah; Akyol, Fadil H.; Atahan, I. Lale

    2002-01-01

    Purpose: To compare the local control and survival rates obtained with either external beam radiation therapy (ERT) and adjuvant high-dose-rate (HDR) brachytherapy (BRT) or ERT alone in patients with nasopharyngeal cancer. Methods and Materials: Between December 1993 and December 1999, 144 patients (106 male, 38 female) with the diagnosis of nasopharyngeal cancer were treated with either ERT and adjuvant HDR BRT (Group A) or ERT alone (Group B) at our department. BRT was not applied in 38 patients for the following reasons: (1) Unit was unavailable (n=13), (2) Patient was younger than 18 years (n=17), (3) Patient received accelerated hyperfractionated ERT (n=6), and (4) Patient refused BRT (n=2). The median age for whole group was 43 (range: 9-82 years). According to the AJCC-1997 staging system, there were 11 (7.6%), 35 (24.3%), 38 (26.4%), and 60 (41.7%) patients in Stage I, II, III, and IV, respectively. There were 57 (39.6%) patients with T1, 41 (28.5%) with T2, 20 (13.9%) with T3, and 26 (18.1%) with T4 tumors. Histopathologic diagnosis was WHO 2-3 in 137 (95.2%) patients. ERT doses ranged between 58.8 and 74 Gy (median: 66 Gy). There were significantly more patients with young age, N2 status, and Stage III disease in Group B and with Stage II disease in Group A. Significantly more patients received chemotherapy in Group B. BRT with an HDR 192 Ir microSelectron afterloading unit was delivered in 106 patients at the conclusion of ERT using a single-channel nasal applicator. Dose was prescribed at 1 cm from the source, and total dose of 12 Gy in 3 fractions on 3 consecutive days was given immediately after ERT. Besides radiotherapy, 82 (56.9%) patients received cisplatin-based chemotherapy, as well. Follow-up time ranged between 12 and 80 months (median: 32 months). Results: The two groups were comparable in terms of local recurrence, locoregional failure, regional failure, and rate of distant metastasis. Local failure was observed in 11 (10.3%) out of 106

  10. Monte Carlo dosimetric characterization of the Flexisource Co-60 high-dose-rate brachytherapy source using PENELOPE.

    Science.gov (United States)

    Almansa, Julio F; Guerrero, Rafael; Torres, Javier; Lallena, Antonio M

    60 Co sources have been commercialized as an alternative to 192 Ir sources for high-dose-rate (HDR) brachytherapy. One of them is the Flexisource Co-60 HDR source manufactured by Elekta. The only available dosimetric characterization of this source is that of Vijande et al. [J Contemp Brachytherapy 2012; 4:34-44], whose results were not included in the AAPM/ESTRO consensus document. In that work, the dosimetric quantities were calculated as averages of the results obtained with the Geant4 and PENELOPE Monte Carlo (MC) codes, though for other sources, significant differences have been quoted between the values obtained with these two codes. The aim of this work is to perform the dosimetric characterization of the Flexisource Co-60 HDR source using PENELOPE. The MC simulation code PENELOPE (v. 2014) has been used. Following the recommendations of the AAPM/ESTRO report, the radial dose function, the anisotropy function, the air-kerma strength, the dose rate constant, and the absorbed dose rate in water have been calculated. The results we have obtained exceed those of Vijande et al. In particular, the absorbed dose rate constant is ∼0.85% larger. A similar difference is also found in the other dosimetric quantities. The effect of the electrons emitted in the decay of 60 Co, usually neglected in this kind of simulations, is significant up to the distances of 0.25 cm from the source. The systematic and significant differences we have found between PENELOPE results and the average values found by Vijande et al. point out that the dosimetric characterizations carried out with the various MC codes should be provided independently. Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  11. Biological effect of Pulsed Dose Rate brachytherapy with stepping sources

    International Nuclear Information System (INIS)

    Limbergen, Erik F.M. van; Fowler, Jack F.

    1996-01-01

    Purpose: To explore the possible increase of radiation effect in tissues irradiated by pulsed brachytherapy (PDR), for local tissue dose-rates between those 'averaged over the whole pulse' and the instantaneous high dose rates close to the dwell positions. An earlier publication (Fowler and Mount 1992) had shown that, for dose rates (averaged for the duration of the pulse) up to 3 Gy/h, little change of isoeffect doses from continuous low dose rate (CLDR) are expected, unless larger doses per fraction than 1 Gy are used, and especially if components of very rapid repair are present with half-times of less than about 0.5 hours. However, local and transient dose rates close to stepping sources can be up to several Gy per minute. Methods: Calculations were done assuming the linear quadratic formula for radiation damage, in which only the dose-squared term is subject to repair, at a constant exponential rate. The formula developed by Dale for fractionated low-dose-rate radiotherapy was used. A constant overall time of 140 hours and constant total dose of 70 Gy were assumed throughout, the continuous low dose-rate of 0.5 Gy/h (CLDR) providing the unitary standard effects for each PDR condition. Effects of dose-rates ranging from 4 Gy/h to 120 Gy/h (HDR at 2 Gy/min) were studied, and T (1(2)) from 4 minutes to 1.5 hours. Results: Curves are presented relating the ratio of increased biological effect (proportional to log cell kill) calculated for PDR relative to CLDR. Ratios as high as 1.5 can be found for large doses per pulse (> 1 Gy) at high instantaneous dose-rates if T (1(2)) in tissues is as short as a few minutes. The major influences on effect are dose per pulse, half-time of repair in the tissue, and - when T (1(2)) is short - the instantaneous dose-rate. Maximum ratios of PDR/CLDR effect occur when the dose-rate is such that pulse duration is approximately equal to T (1(2)) of repair. Results are presented for late-responding tissues, the differences from CLDR

  12. High-dose-rate brachytherapy in the treatment of uterine cervix cancer. Analysis of dose effectiveness and late complications

    International Nuclear Information System (INIS)

    Ferrigno, Robson; Novaes, Paulo Eduardo Ribeiro dos Santos; Pellizzon, Antonio Cassio Assis; Maia, Maria Aparecida Conte; Fogarolli, Ricardo Cesar; Gentil, Andre Cavalcanti; Salvajoli, Joao Victor

    2001-01-01

    Purpose: This retrospective analysis aims to report results of patients with cervix cancer treated by external beam radiotherapy (EBR) and high-dose-rate (HDR) brachytherapy. Methods and Materials: From September 1992 to December 1996, 138 patients with FIGO Stages II and III and mean age of 56 years were treated. Median EBR to the whole pelvis was 45 Gy in 25 fractions. Parametrial boost was performed in 93% of patients, with a median dose of 14.4 Gy. Brachytherapy with HDR was performed during EBR or following its completion with a dose of 24 Gy in four weekly fractions of 6 Gy to point A. Median overall treatment time was of 60 days. Patient age, tumor stage, and overall treatment time were variables analyzed for survival and local control. Cumulative biologic effective dose (BED) at rectal and bladder reference points were correlated with late complications in these organs and dose of EBR at parametrium was correlated with small bowel complications. Results: Median follow-up time was 38 months. Overall survival, disease-free survival, and local control at 5 years was 53.7%, 52.7%, and 62%, respectively. By multivariate and univariate analysis, overall treatment time up to 50 days was the only statistically significant adverse variable for overall survival (p=0.003) and actuarial local control (p=0.008). The 5-year actuarial incidence of rectal, bladder, and small bowel late complications was 16%, 11%, and 14%, respectively. Patients treated with cumulative BED at rectum points above 110 Gy 3 and at bladder point above 125 Gy 3 had a higher but not statistically significant 5-year actuarial rate of complications at these organs (18% vs. 12%, p=0.49 and 17% vs. 9%, p=0.20, respectively). Patients who received parametrial doses larger than 59 Gy had a higher 5-year actuarial rate of complications in the small bowel; however, this was not statistically significant (19% vs. 10%, p=0.260). Conclusion: This series suggests that 45 Gy to the whole pelvis combined with

  13. IPIP: A new approach to inverse planning for HDR brachytherapy by directly optimizing dosimetric indices

    International Nuclear Information System (INIS)

    Siauw, Timmy; Cunha, Adam; Atamtuerk, Alper; Hsu, I-Chow; Pouliot, Jean; Goldberg, Ken

    2011-01-01

    Purpose: Many planning methods for high dose rate (HDR) brachytherapy require an iterative approach. A set of computational parameters are hypothesized that will give a dose plan that meets dosimetric criteria. A dose plan is computed using these parameters, and if any dosimetric criteria are not met, the process is iterated until a suitable dose plan is found. In this way, the dose distribution is controlled by abstract parameters. The purpose of this study is to develop a new approach for HDR brachytherapy by directly optimizing the dose distribution based on dosimetric criteria. Methods: The authors developed inverse planning by integer program (IPIP), an optimization model for computing HDR brachytherapy dose plans and a fast heuristic for it. They used their heuristic to compute dose plans for 20 anonymized prostate cancer image data sets from patients previously treated at their clinic database. Dosimetry was evaluated and compared to dosimetric criteria. Results: Dose plans computed from IPIP satisfied all given dosimetric criteria for the target and healthy tissue after a single iteration. The average target coverage was 95%. The average computation time for IPIP was 30.1 s on an Intel(R) Core TM 2 Duo CPU 1.67 GHz processor with 3 Gib RAM. Conclusions: IPIP is an HDR brachytherapy planning system that directly incorporates dosimetric criteria. The authors have demonstrated that IPIP has clinically acceptable performance for the prostate cases and dosimetric criteria used in this study, in both dosimetry and runtime. Further study is required to determine if IPIP performs well for a more general group of patients and dosimetric criteria, including other cancer sites such as GYN.

  14. Quality assurance of HDR 192Ir sources using a Fricke dosimeter.

    Science.gov (United States)

    Austerlitz, C; Mota, H; Almeida, C E; Allison, R; Sibata, C

    2007-04-01

    A prototype of a Fricke dosimetry system consisting of a 15 x 15 x 15 cm3 water phantom made of Plexiglas and a 11.3-ml Pyrex balloon fitted with a 0.2 cm thick Pyrex sleeve in its center was created to assess source strength and treatment planning algorithms for use in high dose rate (HDR) 192Ir afterloading units. In routine operation, the radioactive source is positioned at the end of a sleeve, which coincides with the center of the spherical balloon that is filled with Fricke solution, so that the solution is nearly isotropically irradiated. The Fricke system was calibrated in terms of source strength against a reference well-type ionization chamber, and in terms of radial dose by means of an existing algorithm from the HDR's treatment planning system. Because the system is based on the Fricke dosimeter itself, for a given type and model of 192Ir source, the system needs initial calibration but no recalibration. The results from measurements made over a 10 month period, including source decay and source substitutions, have shown the feasibility of using such a system for quality control (QC) of HDR afterloading equipment, including both the source activity and treatment planning parameters. The benefit of a large scale production and the use of this device for clinical HDR QC audits via mail are also discussed.

  15. High dose rate brachytherapy for the treatment of soft tissue sarcoma of the extremity

    International Nuclear Information System (INIS)

    Speight, J.L.; Streeter, O.E.; Chawla, S.; Menendez, L.E.

    1996-01-01

    Purpose: we examined the role of preoperative neoadjuvant chemoradiation and adjuvant high-dose rate brachytherapy on the management of prognostically unfavorable soft tissue sarcomas of the extremities. Our goal was to examine the effect of high dose rate interstitial brachytherapy (HDR IBT) on reducing the risk of local recurrence following limb-sparing resection, as well as shortening treatment duration. Materials and methods: eleven patients, ranging in age from 31 to 73 years old, with soft tissue sarcoma of the extremity were treated at USC/Norris Comprehensive Cancer Center during 1994 and 1995. All patients had biopsy proven soft tissue sarcoma, and all were suitable candidates for limb-sparing surgery. All lesions were greater than 5cm in size and were primarily high grade. Tumor histologies included malignant fibrous histiocytoma (45%), liposarcoma (18%) and leiomyosarcoma, synovial cell sarcoma and spindle cell sarcoma (36%). Sites of tumor origin were the lower extremity (55%), upper extremity (18%) and buttock (9%), 1 patient (9%) had lesions in both the upper and lower extremity. Patients received HDR IBT following combined chemotherapy and external beam irradiation (EBRT) and en bloc resection of the sarcoma. Neoadjuvant chemotherapy consisted of three to four cycles of either Ifosfamide/Mesna with or without Adriamycin, or Mesna, Adriamycin, Ifosfamide and Dacarbazine. One patient received Cis-platin in addition to Ifos/Adr. A minimum of two cycles of chemotherapy were administered prior to EBRT. Additional cycles of chemotherapy were completed concurrently with EBRT but prior to HDR IBT. Preoperative EBRT doses ranging from 40 to 59.4 Gy were given in daily fractions of 180 to 200cGy. Following en bloc resection, HDR IBT was administered using the Omnitron tm 2000 remote afterloading system. Doses ranging from 13 to 30 Gy were delivered to the surgical tumor bed at depths of 0.5mm to 0.75mm from the radioactive source. Results: median follow-up was

  16. Dose and dose rate effects of whole-body gamma-irradiation: II. Hematological variables and cytokines

    Science.gov (United States)

    Gridley, D. S.; Pecaut, M. J.; Miller, G. M.; Moyers, M. F.; Nelson, G. A.

    2001-01-01

    The goal of part II of this study was to evaluate the effects of gamma-radiation on circulating blood cells, functional characteristics of splenocytes, and cytokine expression after whole-body irradiation at varying total doses and at low- and high-dose-rates (LDR, HDR). Young adult C57BL/6 mice (n = 75) were irradiated with either 1 cGy/min or 80 cGy/min photons from a 60Co source to cumulative doses of 0.5, 1.5, and 3.0 Gy. The animals were euthanized at 4 days post-exposure for in vitro assays. Significant dose- (but not dose-rate-) dependent decreases were observed in erythrocyte and blood leukocyte counts, hemoglobin, hematocrit, lipopolysaccharide (LPS)-induced 3H-thymidine incorporation, and interleukin-2 (IL-2) secretion by activated spleen cells when compared to sham-irradiated controls (p factor-beta 1 (TGF-beta 1) and splenocyte secretion of tumor necrosis factor-alpha (TNF-alpha) were not affected by either the dose or dose rate of radiation. The data demonstrate that the responses of blood and spleen were largely dependent upon the total dose of radiation employed and that an 80-fold difference in the dose rate was not a significant factor in the great majority of measurements.

  17. γ irradiation with different dose rates induces different DNA damage responses in Petunia x hybrida cells.

    Science.gov (United States)

    Donà, Mattia; Ventura, Lorenzo; Macovei, Anca; Confalonieri, Massimo; Savio, Monica; Giovannini, Annalisa; Carbonera, Daniela; Balestrazzi, Alma

    2013-05-15

    In plants, there is evidence that different dose rate exposures to gamma (γ) rays can cause different biological effects. The dynamics of DNA damage accumulation and molecular mechanisms that regulate recovery from radiation injury as a function of dose rate are poorly explored. To highlight dose-rate dependent differences in DNA damage, single cell gel electrophoresis was carried out on regenerating Petunia x hybrida leaf discs exposed to LDR (total dose 50 Gy, delivered at 0.33 Gy min(-1)) and HDR (total doses 50 and 100 Gy, delivered at 5.15 Gy min(-1)) γ-ray in the 0-24h time period after treatments. Significant fluctuations of double strand breaks and different repair capacities were observed between treatments in the 0-4h time period following irradiation. Dose-rate-dependent changes in the expression of the PhMT2 and PhAPX genes encoding a type 2 metallothionein and the cytosolic isoform of ascorbate peroxidase, respectively, were detected by Quantitative RealTime-Polymerase Chain Reaction. The PhMT2 and PhAPX genes were significantly up-regulated (3.0- and 0.7-fold) in response to HDR. The results are discussed in light of the potential practical applications of LDR-based treatments in mutation breeding. Copyright © 2013 Elsevier GmbH. All rights reserved.

  18. A quality indicator to evaluate high-dose-rate intracavitary brachytherapy for cancer of the cervix

    International Nuclear Information System (INIS)

    Morales, Francisco Contreras; Soboll, Daniel Scheidegger

    2000-01-01

    The aim of this report is to prevent a simple quality indicator (QI) that can be promptly used to evaluate the high-dose-rate (HDR) intracavitary brachytherapy for the treatment of cancer of the cervix, and if necessary, to correct applicators' geometry before starting the treatment. We selected 51 HDR intracavitary applications of brachytherapy of patients with carcinoma of the cervix treated with 60 mm uterine tandem and small Fletcher colpostat, according to the Manchester method (dose prescription on point A). A QI was defined as the ratio between the volume of 100% isodose curve of the study insertion and the volume of the 100% isodose curve of an insertion considered to be ideal. The data obtained were distributed in three groups: the group with tandem placement slippage (67,5%), a group with colpostat placement slippage (21,9%), and a third group, considered normal (10,6%). Each group showed particular characteristics (p < 0.0001). QI can be the best auxiliary method to establish the error tolerance (%) allowed for HDR intracavitary brachytherapy. (author)

  19. High-Dose-Rate Brachytherapy Boost for Prostate Cancer: Comparison of Two Different Fractionation Schemes

    International Nuclear Information System (INIS)

    Kaprealian, Tania; Weinberg, Vivian; Speight, Joycelyn L.; Gottschalk, Alexander R.; Roach, Mack; Shinohara, Katsuto; Hsu, I.-Chow

    2012-01-01

    Purpose: This is a retrospective study comparing our experience with high-dose-rate (HDR) brachytherapy boost for prostate cancer, using two different fractionation schemes, 600 cGy × 3 fractions (patient group 1) and 950 cGy × 2 fractions (patient group 2). Methods and Materials: A total of 165 patients were treated for prostate cancer using external beam radiation therapy up to a dose of 45 Gy, followed by an HDR brachytherapy prostate radiation boost. Between July 1997 and Nov 1999, 64 patients were treated with an HDR boost of 600 cGy × 3 fractions; and between June 2000 and Nov 2005, 101 patients were treated with an HDR boost of 950 cGy × 2 fractions. All but 9 patients had at least one of the following risk features: pretreatment prostate-specific antigen (PSA) level >10, a Gleason score ≥7, and/or clinical stage T3 disease. Results: Median follow-up was 105 months for group 1 and 43 months for group 2. Patients in group 2 had a greater number of high-risk features than group 1 (p = 0.02). Adjusted for comparable follow-up, there was no difference in biochemical no-evidence-of-disease (bNED) rate between the two fractionation scheme approaches, with 5-year Kaplan-Meier estimates of 93.5% in group 1 and 87.3% in group 2 (p = 0.19). The 5-year estimates of progression-free survival were 86% for group 1 and 83% for group 2 (p = 0.53). Among high-risk patients, there were no differences in bNED or PFS rate due to fractionation. Conclusions: Results were excellent for both groups. Adjusted for comparable follow-up, no differences were found between groups.

  20. Multiobjective anatomy-based dose optimization for HDR-brachytherapy with constraint free deterministic algorithms

    International Nuclear Information System (INIS)

    Milickovic, N.; Lahanas, M.; Papagiannopoulou, M.; Zamboglou, N.; Baltas, D.

    2002-01-01

    In high dose rate (HDR) brachytherapy, conventional dose optimization algorithms consider multiple objectives in the form of an aggregate function that transforms the multiobjective problem into a single-objective problem. As a result, there is a loss of information on the available alternative possible solutions. This method assumes that the treatment planner exactly understands the correlation between competing objectives and knows the physical constraints. This knowledge is provided by the Pareto trade-off set obtained by single-objective optimization algorithms with a repeated optimization with different importance vectors. A mapping technique avoids non-feasible solutions with negative dwell weights and allows the use of constraint free gradient-based deterministic algorithms. We compare various such algorithms and methods which could improve their performance. This finally allows us to generate a large number of solutions in a few minutes. We use objectives expressed in terms of dose variances obtained from a few hundred sampling points in the planning target volume (PTV) and in organs at risk (OAR). We compare two- to four-dimensional Pareto fronts obtained with the deterministic algorithms and with a fast-simulated annealing algorithm. For PTV-based objectives, due to the convex objective functions, the obtained solutions are global optimal. If OARs are included, then the solutions found are also global optimal, although local minima may be present as suggested. (author)

  1. Comparison of the outcome and morbidity for localized or locally advanced prostate cancer treated by high-dose-rate brachytherapy plus external beam radiotherapy (EBRT) versus EBRT alone

    International Nuclear Information System (INIS)

    Fang Fumin; Wang Yuming; Wang Chongjong; Huang Hsuanying; Chiang Pohui

    2008-01-01

    The objective of this study was to compare the survival, gastrointestinal (GI) and genitourinary (GU) toxicity for localized or locally advanced prostate cancer treated by high-dose-rate-brachytherapy (HDR-BT) plus external beam radiotherapy (EBRT) versus EBRT alone at a single institute in Taiwan. Eighty-eight patients with T1c-T3b prostate cancer consecutively treated by EBRT alone (33 patients) or HDR-BT+EBRT (55 patients) were studied. The median dose of EBRT was 70.2 Gy in the EBRT group and 50.4 Gy in the HDR-BT group. HDR-BT was performed 2-3 weeks before EBRT, with 12.6 Gy in three fractions over 24 h. Five patients (15.2%) in the EBRT group and seven (12.7%) in the HDR-BT group developed a biochemical relapse. The 5-year actuarial biochemical relapse-free survival rates were 65.0% in the EBRT group and 66.7% in the HDR-BT group (P=0.76). The 5-year actuarial likelihood of late ≥Grade 2 and ≥Grade 3 GI toxicity in the EBRT versus HDR-BT group was 62.8 versus 7.7% (P<0.001) and 19.6 versus 0% (P=0.001), respectively. In a multivariate analysis, the only predictor for late GI toxicity was the mode of RT. The 5-year actuarial likelihood of late ≥Grade 2 and ≥Grade 3 GU toxicity in the EBRT versus HDR-BT group was 14.8 versus 15.9% (P=0.86) and 3.6 versus 8.5% (P=0.40), respectively. The addition of HDR-BT before EBRT with a reduced dose from the EBRT produces a comparable survival outcome and GU toxicity but a significantly less GI toxicity for prostate cancer patients. (author)

  2. A generic high-dose rate {sup 192}Ir brachytherapy source for evaluation of model-based dose calculations beyond the TG-43 formalism

    Energy Technology Data Exchange (ETDEWEB)

    Ballester, Facundo, E-mail: Facundo.Ballester@uv.es [Department of Atomic, Molecular and Nuclear Physics, University of Valencia, Burjassot 46100 (Spain); Carlsson Tedgren, Åsa [Department of Medical and Health Sciences (IMH), Radiation Physics, Faculty of Health Sciences, Linköping University, Linköping SE-581 85, Sweden and Department of Medical Physics, Karolinska University Hospital, Stockholm SE-171 76 (Sweden); Granero, Domingo [Department of Radiation Physics, ERESA, Hospital General Universitario, Valencia E-46014 (Spain); Haworth, Annette [Department of Physical Sciences, Peter MacCallum Cancer Centre and Royal Melbourne Institute of Technology, Melbourne, Victoria 3000 (Australia); Mourtada, Firas [Department of Radiation Oncology, Helen F. Graham Cancer Center, Christiana Care Health System, Newark, Delaware 19713 (United States); Fonseca, Gabriel Paiva [Instituto de Pesquisas Energéticas e Nucleares – IPEN-CNEN/SP, São Paulo 05508-000, Brazil and Department of Radiation Oncology (MAASTRO), GROW, School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN (Netherlands); Zourari, Kyveli; Papagiannis, Panagiotis [Medical Physics Laboratory, Medical School, University of Athens, 75 MikrasAsias, Athens 115 27 (Greece); Rivard, Mark J. [Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111 (United States); Siebert, Frank-André [Clinic of Radiotherapy, University Hospital of Schleswig-Holstein, Campus Kiel, Kiel 24105 (Germany); Sloboda, Ron S. [Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta T6G 1Z2, Canada and Department of Oncology, University of Alberta, Edmonton, Alberta T6G 2R3 (Canada); and others

    2015-06-15

    Purpose: In order to facilitate a smooth transition for brachytherapy dose calculations from the American Association of Physicists in Medicine (AAPM) Task Group No. 43 (TG-43) formalism to model-based dose calculation algorithms (MBDCAs), treatment planning systems (TPSs) using a MBDCA require a set of well-defined test case plans characterized by Monte Carlo (MC) methods. This also permits direct dose comparison to TG-43 reference data. Such test case plans should be made available for use in the software commissioning process performed by clinical end users. To this end, a hypothetical, generic high-dose rate (HDR) {sup 192}Ir source and a virtual water phantom were designed, which can be imported into a TPS. Methods: A hypothetical, generic HDR {sup 192}Ir source was designed based on commercially available sources as well as a virtual, cubic water phantom that can be imported into any TPS in DICOM format. The dose distribution of the generic {sup 192}Ir source when placed at the center of the cubic phantom, and away from the center under altered scatter conditions, was evaluated using two commercial MBDCAs [Oncentra{sup ®} Brachy with advanced collapsed-cone engine (ACE) and BrachyVision ACUROS{sup TM}]. Dose comparisons were performed using state-of-the-art MC codes for radiation transport, including ALGEBRA, BrachyDose, GEANT4, MCNP5, MCNP6, and PENELOPE2008. The methodologies adhered to recommendations in the AAPM TG-229 report on high-energy brachytherapy source dosimetry. TG-43 dosimetry parameters, an along-away dose-rate table, and primary and scatter separated (PSS) data were obtained. The virtual water phantom of (201){sup 3} voxels (1 mm sides) was used to evaluate the calculated dose distributions. Two test case plans involving a single position of the generic HDR {sup 192}Ir source in this phantom were prepared: (i) source centered in the phantom and (ii) source displaced 7 cm laterally from the center. Datasets were independently produced by

  3. Clinical outcome of high-dose-rate interstitial brachytherapy in patients with oral cavity cancer

    International Nuclear Information System (INIS)

    Lee, Sung Uk; Cho, Kwan Ho; Moon, Sung Ho; Choi, Sung Weon; Park, Joo Yong; Yun, Tak; Lee, Sang Hyun; Lim, Young Kyung; Jeong, Chi Young

    2014-01-01

    To evaluate the clinical outcome of high-dose-rate (HDR) interstitial brachytherapy (IBT) in patients with oral cavity cancer. Sixteen patients with oral cavity cancer treated with HDR remote-control afterloading brachytherapy using 192Ir between 2001 and 2013 were analyzed retrospectively. Brachytherapy was administered in 11 patients as the primary treatment and in five patients as salvage treatment for recurrence after the initial surgery. In 12 patients, external beam radiotherapy (50-55 Gy/25 fractions) was combined with IBT of 21 Gy/7 fractions. In addition, IBT was administered as the sole treatment in three patients with a total dose of 50 Gy/10 fractions and as postoperative adjuvant treatment in one patient with a total of 35 Gy/7 fractions. The 5-year overall survival of the entire group was 70%. The actuarial local control rate after 3 years was 84%. All five recurrent cases after initial surgery were successfully salvaged using IBT +/- external beam radiotherapy. Two patients developed local recurrence at 3 and 5 months, respectively, after IBT. The acute complications were acceptable (< or =grade 2). Three patients developed major late complications, such as radio-osteonecrosis, in which one patient was treated by conservative therapy and two required surgical intervention. HDR IBT for oral cavity cancer was effective and acceptable in diverse clinical settings, such as in the cases of primary or salvage treatment.

  4. Significance of manipulating tumour hypoxia and radiation dose rate in terms of local tumour response and lung metastatic potential, referring to the response of quiescent cell populations

    Science.gov (United States)

    Masunaga, S; Matsumoto, Y; Kashino, G; Hirayama, R; Liu, Y; Tanaka, H; Sakurai, Y; Suzuki, M; Kinashi, Y; Maruhashi, A; Ono, K

    2010-01-01

    The purpose of this study was to evaluate the influence of manipulating intratumour oxygenation status and radiation dose rate on local tumour response and lung metastases following radiotherapy, referring to the response of quiescent cell populations within irradiated tumours. B16-BL6 melanoma tumour-bearing C57BL/6 mice were continuously given 5-bromo-2′-deoxyuridine (BrdU) to label all proliferating (P) cells. They received γ-ray irradiation at high dose rate (HDR) or reduced dose rate (RDR) following treatment with the acute hypoxia-releasing agent nicotinamide or local hyperthermia at mild temperatures (MTH). Immediately after the irradiation, cells from some tumours were isolated and incubated with a cytokinesis blocker. The responses of the quiescent (Q) and total (proliferating + Q) cell populations were assessed based on the frequency of micronuclei using immunofluorescence staining for BrdU. In other tumour-bearing mice, 17 days after irradiation, macroscopic lung metastases were enumerated. Following HDR irradiation, nicotinamide and MTH enhanced the sensitivity of the total and Q-cell populations, respectively. The decrease in sensitivity at RDR irradiation compared with HDR irradiation was slightly inhibited by MTH, especially in Q cells. Without γ-ray irradiation, nicotinamide treatment tended to reduce the number of lung metastases. With γ-rays, in combination with nicotinamide or MTH, especially the former, HDR irradiation decreased the number of metastases more remarkably than RDR irradiation. Manipulating both tumour hypoxia and irradiation dose rate have the potential to influence lung metastasis. The combination with the acute hypoxia-releasing agent nicotinamide may be more promising in HDR than RDR irradiation in terms of reducing the number of lung metastases. PMID:20739345

  5. High dose rate brachytherapy using custom made superficial mould applicators and Leipzig applicators for non melanoma localized skin cancer

    Energy Technology Data Exchange (ETDEWEB)

    Pellizzon, A. Cassio A.; Miziara, Daniela; Lima, Flavia Pedroso de; Miziara, Miguel

    2014-07-01

    Purpose: advances in technology and the commercial production of Leipzig applicators allowed High Dose Rate after-load brachytherapy (HDR-BT) to address a number of the challenges associated with the delivery of superficial radiation to treat localized non melanoma skin cancer (NMSK). We reviewed our uni-institutional experience on the treatment of NMSK with HDR-BT. Methods: data were collected retrospectively from patients attending the Radiation Oncology Department at AV Carvalho Insitute, Sao Paulo, Brazil. HDR-BT was done using the stepping source HDR 192Ir Microselectron (Nucletron BV). The planning target volume consisted of the macroscopic lesion plus a 5mm to 10mm margin.The depth of treatment was 0.5 cm in smaller (< 2.0 cm) tumors and 10 to 15 mm for lesions bigger than that. Results: Thirteen patients were treated with HDR-BT from June, 2007 to June 2013. The median age and follow up time were 72 (38-90) years old and 36 (range, 7-73) months, respectively. There a predominance of males (61.5%) and of patients referred for adjuvant treatment due positive surgical margins or because they have had only a excision biopsy without safety margins (61.5%). Six (46.2%) patients presented with squamous cell carcinoma and 7 (53.8%) patients presented with basal cell carcinoma. The median tumor size was 20 (range, 5-42) mm. Patients were treated with a median total dose of 40 Gy (range, 20 -60), given in 10 (range, 2-15) fractions, given daily or twice a week. All patients responded very well to treatment and only one patient has failed locally so far, after 38 months of the end of the irradiation. The crude and actuarial 3-year local control rates were 100% and 80%, respectively. Moist desquamation, grade 2 RTOG, was observed in 4 (30.8%) patients. Severe late complication, radiation-induced dyspigmentation, occurred in 2 patients and 1 of the patients also showed telangiectasia in the irradiated area. The cosmetic result was considered good in 84% (11/13) patients

  6. High dose rate brachytherapy using custom made superficial mould applicators and Leipzig applicators for non melanoma localized skin cancer

    International Nuclear Information System (INIS)

    Pellizzon, A. Cassio A.; Miziara, Daniela; Lima, Flavia Pedroso de; Miziara, Miguel

    2014-01-01

    Purpose: advances in technology and the commercial production of Leipzig applicators allowed High Dose Rate after-load brachytherapy (HDR-BT) to address a number of the challenges associated with the delivery of superficial radiation to treat localized non melanoma skin cancer (NMSK). We reviewed our uni-institutional experience on the treatment of NMSK with HDR-BT. Methods: data were collected retrospectively from patients attending the Radiation Oncology Department at AV Carvalho Insitute, Sao Paulo, Brazil. HDR-BT was done using the stepping source HDR 192Ir Microselectron (Nucletron BV). The planning target volume consisted of the macroscopic lesion plus a 5mm to 10mm margin.The depth of treatment was 0.5 cm in smaller (< 2.0 cm) tumors and 10 to 15 mm for lesions bigger than that. Results: Thirteen patients were treated with HDR-BT from June, 2007 to June 2013. The median age and follow up time were 72 (38-90) years old and 36 (range, 7-73) months, respectively. There a predominance of males (61.5%) and of patients referred for adjuvant treatment due positive surgical margins or because they have had only a excision biopsy without safety margins (61.5%). Six (46.2%) patients presented with squamous cell carcinoma and 7 (53.8%) patients presented with basal cell carcinoma. The median tumor size was 20 (range, 5-42) mm. Patients were treated with a median total dose of 40 Gy (range, 20 -60), given in 10 (range, 2-15) fractions, given daily or twice a week. All patients responded very well to treatment and only one patient has failed locally so far, after 38 months of the end of the irradiation. The crude and actuarial 3-year local control rates were 100% and 80%, respectively. Moist desquamation, grade 2 RTOG, was observed in 4 (30.8%) patients. Severe late complication, radiation-induced dyspigmentation, occurred in 2 patients and 1 of the patients also showed telangiectasia in the irradiated area. The cosmetic result was considered good in 84% (11/13) patients

  7. Phase I/II trial of single-fraction high-dose-rate brachytherapy-boosted hypofractionated intensity-modulated radiation therapy for localized adenocarcinoma of the prostate.

    Science.gov (United States)

    Myers, Michael A; Hagan, Michael P; Todor, Dorin; Gilbert, Lynn; Mukhopadhyay, Nitai; Randolf, Jessica; Heimiller, Jeffrey; Anscher, Mitchell S

    2012-01-01

    A Phase I/II protocol was conducted to examine the toxicity and efficacy of the combination of intensity-modulated radiation therapy (IMRT) with a single-fraction high-dose-rate (HDR) brachytherapy implant. From 2001 through 2006, 26 consecutive patients were treated on the trial. The primary objective was to demonstrate a high rate of completion without experiencing a treatment-limiting toxicity. Eligibility was limited to patients with T stage ≤2b, prostate-specific antigen (PSA) ≤20, and Gleason score ≤7. Treatment began with a single HDR fraction of 6Gy to the entire prostate and 9Gy to the peripheral zone, followed by IMRT optimized to deliver in 28 fractions with a normalized total dose of 70Gy. Patients received 50.4Gy to the pelvic lymph node. The prostate dose (IMRT and HDR) resulted in an average biologic equivalent dose >128Gy (α/β=3). Patients whose pretreatment PSA was ≥10ng/mL, Gleason score 7, or stage ≥T2b received short-term androgen ablation. Median followup was 53 months (9-68 months). There were no biochemical failures by either the American Society of Therapeutic Radiology and Oncology or the Phoenix definitions. The median nadir PSA was 0.32ng/mL. All the 26 patients completed the treatment as prescribed. The rate of Grade 3 late genitourinary toxicity was 3.8% consisting of a urethral stricture. There was no other Grade 3 or 4 genitourinary or gastrointestinal toxicities. Single-fraction HDR-boosted IMRT is a safe effective method of dose escalation for localized prostate cancer. Copyright © 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  8. Dose and dose rate effects of whole-body gamma-irradiation: II. Hematological variables and cytokines

    Science.gov (United States)

    Gridley, D. S.; Pecaut, M. J.; Miller, G. M.; Moyers, M. F.; Nelson, G. A.

    2001-01-01

    The goal of part II of this study was to evaluate the effects of gamma-radiation on circulating blood cells, functional characteristics of splenocytes, and cytokine expression after whole-body irradiation at varying total doses and at low- and high-dose-rates (LDR, HDR). Young adult C57BL/6 mice (n = 75) were irradiated with either 1 cGy/min or 80 cGy/min photons from a 60Co source to cumulative doses of 0.5, 1.5, and 3.0 Gy. The animals were euthanized at 4 days post-exposure for in vitro assays. Significant dose- (but not dose-rate-) dependent decreases were observed in erythrocyte and blood leukocyte counts, hemoglobin, hematocrit, lipopolysaccharide (LPS)-induced 3H-thymidine incorporation, and interleukin-2 (IL-2) secretion by activated spleen cells when compared to sham-irradiated controls (p < 0.05). Basal proliferation of leukocytes in the blood and spleen increased significantly with increasing dose (p < 0.05). Significant dose rate effects were observed only in thrombocyte counts. Plasma levels of transforming growth factor-beta 1 (TGF-beta 1) and splenocyte secretion of tumor necrosis factor-alpha (TNF-alpha) were not affected by either the dose or dose rate of radiation. The data demonstrate that the responses of blood and spleen were largely dependent upon the total dose of radiation employed and that an 80-fold difference in the dose rate was not a significant factor in the great majority of measurements.

  9. Essential results of analyses accompanying the leak rate experiments E22 at HDR

    International Nuclear Information System (INIS)

    Grebner, H.; Hoefler, A.; Hunger, H.

    1994-01-01

    Under the E22 test group of phase III of the HDR safety programme, experiments were performed on the crack opening and leak rate behaviour of pipe components of smaller nominal bores. The experiments were complemented by computations, in particular verifications, to qualify the computation models as one of the main aims of the HDR safety programme. Most of the analyses to determine crack openings were performed by means of the finite-element method, including elastic-plastic materials behaviour and, complementarily, assessing engineering methods. The leak rate was calculated by means of separate 2-phase computation models. Altogether, it may be concluded from the structural and fracture mechanical experiments with pipes, elbows and branch pieces, that crack openings and incipient cracks at loading with internal pressure or bending moment can be described with good accuracy by means of the finite-element programme ADINA and the developed FE-models. (orig.) [de

  10. Adherence to Vaginal Dilation Following High Dose Rate Brachytherapy for Endometrial Cancer

    International Nuclear Information System (INIS)

    Friedman, Lois C.; Abdallah, Rita; Schluchter, Mark; Panneerselvam, Ashok; Kunos, Charles A.

    2011-01-01

    Purpose: We report demographic, clinical, and psychosocial factors associated with adherence to vaginal dilation and describe the sexual and marital or nonmarital dyadic functioning of women following high dose rate (HDR) brachytherapy for endometrial cancer. Methods and Materials: We retrospectively evaluated women aged 18 years or older in whom early-stage endometrial (IAgr3-IIB) cancers were treated by HDR intravaginal brachytherapy within the past 3.5 years. Women with or without a sexual partner were eligible. Patients completed questionnaires by mail or by telephone assessing demographic and clinical variables, adherence to vaginal dilation, dyadic satisfaction, sexual functioning, and health beliefs. Results: Seventy-eight of 89 (88%) eligible women with early-stage endometrial cancer treated with HDR brachytherapy completed questionnaires. Only 33% of patients were adherers, based on reporting having used a dilator more than two times per week in the first month following radiation. Nonadherers who reported a perceived change in vaginal dimension following radiation reported that their vaginas were subjectively smaller after brachytherapy (p = 0.013). Adherers reported more worry about their sex lives or lack thereof than nonadherers (p = 0.047). Patients reported considerable sexual dysfunction following completion of HDR brachytherapy. Conclusions: Adherence to recommendations for vaginal dilator use following HDR brachytherapy for endometrial cancer is poor. Interventions designed to educate women about dilator use benefit may increase adherence. Although sexual functioning was compromised, it is likely that this existed before having cancer for many women in our study.

  11. 3-D conformal treatment of prostate cancer to 74 Gy vs. high-dose-rate brachytherapy boost: A cross-sectional quality-of-life survey

    International Nuclear Information System (INIS)

    Vordermark, Dirk

    2006-01-01

    The effects of two modalities of dose-escalated radiotherapy on health-related quality of life (HRQOL) were compared. Forty-one consecutive patients were treated with a 3-D conformal (3-DC) boost to 74 Gy, and 43 with high-dose rate (HDR) brachytherapy boost (2x9 Gy), following 3-D conformal treatment to 46 Gy. Median age was 70 years in both groups, median initial PSA was 7.9 μg/l in 3-DC boost patients and 8.1 μg/l in HDR boost patients. Stage was 7 in 52% and 47%, respectively. HRQOL was assessed cross-sectionally using EORTC QLQ-C30 and organ-specific PR25 modules 3-32 (median 19) and 4-25 (median 14) months after treatment, respectively. Questionnaires were completed by 93% and 97% of patients, respectively. Diarrhea and insomnia scores were significantly increased in both groups. In the PR25 module, scores of 3-DC boost and HDR boost patients for urinary, bowel and treatment-related symptoms were similar. Among responders, 34% of 3-DC boost patients and 86% of HDR boost patients had severe erectile problems. Dose escalation in prostate cancer by either 3-DC boost to 74 Gy or HDR brachytherapy boost appears to result in similar HRQOL profiles

  12. Custom-made micro applicators for high-dose-rate brachytherapy treatment of chronic psoriasis

    Directory of Open Access Journals (Sweden)

    Ivan M. Buzurovic

    2017-06-01

    Full Text Available Purpose: In this study, we present the treatment of the psoriatic nail beds of patients refractory to standard therapies using high-dose-rate (HDR brachytherapy. The custom-made micro applicators (CMMA were designed and constructed for radiation dose delivery to small curvy targets with complicated topology. The role of the HDR brachytherapy treatment was to stimulate the T cells for an increased immune response. Material and methods: The patient diagnosed with psoriatic nail beds refractory to standard therapies received monthly subunguinal injections that caused significant pain and discomfort in both hands. The clinical target was defined as the length from the fingertip to the distal interphalangeal joint. For the accurate and reproducible setup in the multi-fractional treatment delivery, the CMMAs were designed. Five needles were embedded into the dense plastic mesh and covered with 5 mm bolus material for each micro applicator. Five CMMAs were designed, resulting in the usage of 25 catheters in total. Results: The prescription dose was planned to the depth of the anterior surface of the distal phalanx, allowing for the sparing of the surrounding tissue. The total number of the active dwell positions was 145 with step size of 5 mm. The total treatment time was 115 seconds with a 7.36 Ci activity of the 192Ir source. The treatment resulted in good pain control. The patient did not require further injections to the nail bed. After this initial treatment, additional two patients with similar symptoms received HDR brachytherapy. The treatment outcome was favorable in all cases. Conclusions : The first HDR brachytherapy treatment of psoriasis of the nail bed is presented. The initial experience revealed that brachytherapy treatment was well-tolerated and resulted in adequate control of the disease. A larger cohort of patients will be required for additional conclusions related to the long-term clinical benefits.

  13. High-dose-rate brachytherapy as salvage modality for locally recurrent prostate cancer after definitive radiotherapy. A systematic review

    International Nuclear Information System (INIS)

    Chatzikonstantinou, Georgios; Zamboglou, Nikolaos; Roedel, Claus; Tselis, Nikolaos; Zoga, Eleni; Strouthos, Iosif; Butt, Saeed Ahmed

    2017-01-01

    To review the current status of interstitial high-dose-rate brachytherapy as a salvage modality (sHDR BRT) for locally recurrent prostate cancer after definitive radiotherapy (RT). A literature search was performed in PubMed using ''high-dose-rate, brachytherapy, prostate cancer, salvage'' as search terms. In all, 51 search results published between 2000 and 2016 were identified. Data tables were generated and summary descriptions created. The main outcome parameters used were biochemical control (BC) and toxicity scores. Eleven publications reported clinical outcome and toxicity with follow-up ranging from 4-191 months. A variety of dose and fractionation schedules were described, including 19.0 Gy in 2 fractions up to 42.0 Gy in 6 fractions. The 5-year BC ranged from 18-77%. Late grade 3 genitourinary and gastrointestinal toxicity was 0-32% and 0-5.1%, respectively. sHDR BRT appears as safe and effective salvage modality for the reirradiation of locally recurrent prostate cancer after definitive RT. (orig.) [de

  14. Management of a HDR brachytherapy system in the Hospital Juarez of Mexico; Gestion de un sistema de braquiterapia HDR een el Hospital Juarez de Mexico

    Energy Technology Data Exchange (ETDEWEB)

    Serrano F, A.G.; Ramirez R, G.; Gil G, R. [Hospital Juarez de Mexico, Av. l.P.N. 5160, Col. Magdalena de las Salinas, 07760 Mexico D.F. (Mexico); Azorin N, J. [UAM-I, 09340 Mexico D.F. (Mexico); Rivera M, T. [Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Unidad Legaria del IPN, Av. Legaria 694, Col. Irrigacion, 11500 Mexico D.F. (Mexico)

    2007-07-01

    Full text: In the Hospital Juarez of Mexico, it is carried out a project to implement a Brachytherapy system with high dose rate (HDR) through a Management quality program. In our work center this treatment modality in patients with cervicouterine cancer is used (CaCu), and constantly it is necessary to carry out improvements in the procedures, with the purpose of optimizing them and in consequence to complete the principles of the Radiological Protection, guaranteeing this way, an attention with the quality and safety, such that allow to diminish the risks to the patients and to assure that the received dose in critical organs it finds inside the permitted therapeutic limits, without commit the radiosensitive response of healthy organs. In this work an analysis of the implementation of this system is presented, detailing the procedures so much in the technological infrastructure like human and indicating the necessary technical and operative requirements to reach an adequate practice in HDR brachytherapy. (Author)

  15. Essential results of analyses accompanying the leak rate experiments E22 at HDR

    International Nuclear Information System (INIS)

    Grebner, H.; Hoefler, A.; Hunger, H.

    1997-01-01

    During phase III of the HDR Safety Programme (HDR: decommissioned overheated steam reactor in Karlstein, Germany), experiments were performed in test group E22 on small-bore austenitic straight piping and on pipe elbows and branches containing through-wall cracks. The main aim was the determination of crack opening and leak rate behaviour for the cracked components under almost operational pressure and temperature loading conditions, especially including transient bending moments. In addition to machined slits, naturally grown fatigue cracks were also considered to cover the leakage behaviour. The experiments were accompanied by calculations, mainly performed by GRS. The paper describes the most important aspects and the essential results of the calculations and analysis. The main outcome was that the crack opening and initiation of crack growth can be described with the finite element techniques applied with sufficient accuracy. However, the qualification of the leak rate models could not be completed successfully, and therefore more sophisticated experiments of this kind are needed. (orig.)

  16. High dose rate 192Ir calibration: Indonesia experiences

    International Nuclear Information System (INIS)

    Nasukha; Tjiptanto, D.; Darmasyah, R.; Kurniawan, B.

    2002-01-01

    Indonesia with a population of more than 200 Million people which spread on about 5000 islands, up to now only has 23 radiotherapy centers and some not active anymore. As mention by Parkin et al that Cervix/Utery and breast cancer are the most estimated numbers of new cases of cancers in women for developing countries, stomach and lung cancers in men. Indonesia as a developing country is likely similar to other developing countries on numbers of new cases of cancers in women. But quite different in men, in Indonesia the most common cancers are nasopharynx and thyroid cancers. The use of lr-192 sources in high dose-rate (HDR) remotely afterloaded brachytherapy treatments have greatly increased in recent years and variety of such sources are commercially available. Nine radiotherapy centers in Indonesia installed Nucletron microSelectron HDR remote afterloader. Based on the data of CiptoMangunkusurno Hospital, Jakarta that the most common cancers are the cervix, breast, nasopharynx and thyroid cancers which of percentage are about 31%, 25 %, 13%, and 6 % respectively. It means that the use of HDR 192 Ir brachytherapy has to be an effective tool in the treatments. Two methods have been studied and applied to calibrate HDR 192 Ir brachytherapy in Indonesia, especially for Nucletron microSelectron HDR 192 lr remote afterloader brachytherapy. Calibration of HDR 192 Ir brachytherapy source has been done by Cavity lonization Chamber and with Well Type lonization Chamber. First, 0.6 cc of NE Farmer type dosimeter that was calibrated to 60 Co and 250 kV of x-rays in air kerma was used in this experiment. Position of measurement (detector and source) at the center of the room and about 1 meter from the floor. Eight variation of distances from 10 cm to 40 cms have been carried out measurement as recommended by IAEA-TECDOC-1079. Correction have been given for scatters, non-uniformity, and attenuation. To solve the problem of scatter correction factor was used Matlab programming

  17. Prostate-specific antigen bounce after high-dose rate brachytherapy with external beam radiation therapy for prostate cancer patients

    International Nuclear Information System (INIS)

    Sakamoto, Naotaka; Kakinoki, Hiroaki; Tsutsui, Akio; Yoshikawa, Masahiro; Iguchi, Atsushi; Matsunobu, Toru; Uehara, Satoru

    2008-01-01

    Prostate-specific antigen (PSA) bounce after high-dose rate (HDR) brachytherapy with external beam radiation therapy (EBRT) for prostate cancer patients was evaluated. Sixty-one patients treated with HDR-brachytherapy followed by EBRT had a minimum follow-up of 12 months (median, 24 months) in our institute. A PSA bounce was defined as a rise of at least 0.1 ng/ml greater than a previous PSA level, with a subsequent decline equal to, or less than, the initial nadir. A PSA bounce was noted in 16 (26.2%) of 61 patients (one patient had a PSA bounce twice). Median time to develop a PSA bounce was 18 months, but 23.5% developed a PSA bounce after 24 months. Median duration of PSA bounce was 6 months and 11.8% had increased PSA within a period of 12 months. Median bounce height was 0.2 ng/ml (range, 0.1 to 3.39 ng/ml). A bounce height of gerater than 2 ng/ml was seen in 11.8%. Clinical characteristics (age, prostate volume, neoadjuvant endocrine therapy, risk classification, stage, pretreatment PSA, Gleason score) do not predict whether or not there will be a PSA bounce. In patients treated with HDR-brachytherapy followed by EBRT, the incidence and characteristics of PSA bounce were similar to those in patients treated with low-dose rate brachytherapy. Physicians should be aware of the possibility of PSA bounce following HDR-brachytherapy with EBRT. (author)

  18. Prostate CT segmentation method based on nonrigid registration in ultrasound-guided CT-based HDR prostate brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Yang, Xiaofeng, E-mail: xyang43@emory.edu; Rossi, Peter; Ogunleye, Tomi; Marcus, David M.; Jani, Ashesh B.; Curran, Walter J.; Liu, Tian [Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia 30322 (United States); Mao, Hui [Department of Radiology and Imaging Sciences, Emory University, Atlanta, Georgia 30322 (United States)

    2014-11-01

    Purpose: The technological advances in real-time ultrasound image guidance for high-dose-rate (HDR) prostate brachytherapy have placed this treatment modality at the forefront of innovation in cancer radiotherapy. Prostate HDR treatment often involves placing the HDR catheters (needles) into the prostate gland under the transrectal ultrasound (TRUS) guidance, then generating a radiation treatment plan based on CT prostate images, and subsequently delivering high dose of radiation through these catheters. The main challenge for this HDR procedure is to accurately segment the prostate volume in the CT images for the radiation treatment planning. In this study, the authors propose a novel approach that integrates the prostate volume from 3D TRUS images into the treatment planning CT images to provide an accurate prostate delineation for prostate HDR treatment. Methods: The authors’ approach requires acquisition of 3D TRUS prostate images in the operating room right after the HDR catheters are inserted, which takes 1–3 min. These TRUS images are used to create prostate contours. The HDR catheters are reconstructed from the intraoperative TRUS and postoperative CT images, and subsequently used as landmarks for the TRUS–CT image fusion. After TRUS–CT fusion, the TRUS-based prostate volume is deformed to the CT images for treatment planning. This method was first validated with a prostate-phantom study. In addition, a pilot study of ten patients undergoing HDR prostate brachytherapy was conducted to test its clinical feasibility. The accuracy of their approach was assessed through the locations of three implanted fiducial (gold) markers, as well as T2-weighted MR prostate images of patients. Results: For the phantom study, the target registration error (TRE) of gold-markers was 0.41 ± 0.11 mm. For the ten patients, the TRE of gold markers was 1.18 ± 0.26 mm; the prostate volume difference between the authors’ approach and the MRI-based volume was 7.28% ± 0

  19. Prostate CT segmentation method based on nonrigid registration in ultrasound-guided CT-based HDR prostate brachytherapy

    Science.gov (United States)

    Yang, Xiaofeng; Rossi, Peter; Ogunleye, Tomi; Marcus, David M.; Jani, Ashesh B.; Mao, Hui; Curran, Walter J.; Liu, Tian

    2014-01-01

    Purpose: The technological advances in real-time ultrasound image guidance for high-dose-rate (HDR) prostate brachytherapy have placed this treatment modality at the forefront of innovation in cancer radiotherapy. Prostate HDR treatment often involves placing the HDR catheters (needles) into the prostate gland under the transrectal ultrasound (TRUS) guidance, then generating a radiation treatment plan based on CT prostate images, and subsequently delivering high dose of radiation through these catheters. The main challenge for this HDR procedure is to accurately segment the prostate volume in the CT images for the radiation treatment planning. In this study, the authors propose a novel approach that integrates the prostate volume from 3D TRUS images into the treatment planning CT images to provide an accurate prostate delineation for prostate HDR treatment. Methods: The authors’ approach requires acquisition of 3D TRUS prostate images in the operating room right after the HDR catheters are inserted, which takes 1–3 min. These TRUS images are used to create prostate contours. The HDR catheters are reconstructed from the intraoperative TRUS and postoperative CT images, and subsequently used as landmarks for the TRUS–CT image fusion. After TRUS–CT fusion, the TRUS-based prostate volume is deformed to the CT images for treatment planning. This method was first validated with a prostate-phantom study. In addition, a pilot study of ten patients undergoing HDR prostate brachytherapy was conducted to test its clinical feasibility. The accuracy of their approach was assessed through the locations of three implanted fiducial (gold) markers, as well as T2-weighted MR prostate images of patients. Results: For the phantom study, the target registration error (TRE) of gold-markers was 0.41 ± 0.11 mm. For the ten patients, the TRE of gold markers was 1.18 ± 0.26 mm; the prostate volume difference between the authors’ approach and the MRI-based volume was 7.28% ± 0

  20. High-Dose-Rate Monotherapy for Localized Prostate Cancer: 10-Year Results

    Energy Technology Data Exchange (ETDEWEB)

    Hauswald, Henrik; Kamrava, Mitchell R.; Fallon, Julia M.; Wang, Pin-Chieh; Park, Sang-June; Van, Thanh; Borja, Lalaine; Steinberg, Michael L.; Demanes, D. Jeffrey, E-mail: JDemanes@mednet.ucla.edu

    2016-03-15

    Purpose: High-dose-rate (HDR) brachytherapy was originally used with external beam radiation therapy (EBRT) to increase the dose to the prostate without injuring the bladder or rectum. Numerous studies have reported HDR brachytherapy is safe and effective. We adapted it for use without EBRT for cases not requiring lymph node treatment. Patients and Methods: We entered the patient demographics, disease characteristics, and treatment parameters into a prospective registry and serially added follow-up data for 448 men with low-risk (n=288) and intermediate-risk (n=160) prostate cancer treated from 1996 to 2009. Their median age was 64 years (range 42-90). The median prostate-specific antigen (PSA) level was 6.0 ng/mL (range 0.2-18.2). The Gleason score was ≤6 in 76% and 7 in 24%. The median dose was 43.5 Gy in 6 fractions. The clinical and biochemical disease control and survival rates were calculated. Adverse events were graded according to the Common Toxicity Criteria of Adverse Events. Results: The median follow-up period was 6.5 years (range 0.3-15.3). The actuarial 6- and 10-year PSA progression-free survival was 98.6% (95% confidence interval [CI] 96.9%-99.4%) and 97.8% (95% CI 95.5%-98.9%). Overall survival at 10 years was 76.7% (95% CI 69.9%-82.2%). The local control, distant metastasis-free survival, and cause-specific survival were 99.7% (95% CI 97.9%-99.9%), 98.9% (95% CI 96.3%-99.7%), and 99.1% (95% CI 95.8%-99.8%). T stage, initial PSA level, Gleason score, National Comprehensive Cancer Network risk group, patient age, and androgen deprivation therapy did not significantly correlate with disease control or survival. No late grade 3 to 4 rectal toxicities developed. Late grade 3 to 4 genitourinary toxicity occurred in 4.9% (grade 3 in 4.7%). Conclusions: HDR monotherapy is a safe and highly effective treatment of low- and intermediate-risk prostate cancer.

  1. Cervical Gross Tumor Volume Dose Predicts Local Control Using Magnetic Resonance Imaging/Diffusion-Weighted Imaging—Guided High-Dose-Rate and Positron Emission Tomography/Computed Tomography—Guided Intensity Modulated Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Dyk, Pawel; Jiang, Naomi; Sun, Baozhou; DeWees, Todd A. [Department of Radiation Oncology, Washington University School of Medicine, St Louis, Missouri (United States); Fowler, Kathryn J.; Narra, Vamsi [Department of Diagnostic Radiology, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, Missouri (United States); Garcia-Ramirez, Jose L.; Schwarz, Julie K. [Department of Radiation Oncology, Washington University School of Medicine, St Louis, Missouri (United States); Grigsby, Perry W., E-mail: pgrigsby@wustl.edu [Department of Radiation Oncology, Washington University School of Medicine, St Louis, Missouri (United States); Division of Nuclear Medicine, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, Missouri (United States); Division of Gynecologic Oncology, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, Missouri (United States); Alvin J. Siteman Cancer Center, Washington University School of Medicine, St Louis, Missouri (United States)

    2014-11-15

    Purpose: Magnetic resonance imaging/diffusion weighted-imaging (MRI/DWI)-guided high-dose-rate (HDR) brachytherapy and {sup 18}F-fluorodeoxyglucose (FDG) — positron emission tomography/computed tomography (PET/CT)-guided intensity modulated radiation therapy (IMRT) for the definitive treatment of cervical cancer is a novel treatment technique. The purpose of this study was to report our analysis of dose-volume parameters predicting gross tumor volume (GTV) control. Methods and Materials: We analyzed the records of 134 patients with International Federation of Gynecology and Obstetrics stages IB1-IVB cervical cancer treated with combined MRI-guided HDR and IMRT from July 2009 to July 2011. IMRT was targeted to the metabolic tumor volume and lymph nodes by use of FDG-PET/CT simulation. The GTV for each HDR fraction was delineated by use of T2-weighted or apparent diffusion coefficient maps from diffusion-weighted sequences. The D100, D90, and Dmean delivered to the GTV from HDR and IMRT were summed to EQD2. Results: One hundred twenty-five patients received all irradiation treatment as planned, and 9 did not complete treatment. All 134 patients are included in this analysis. Treatment failure in the cervix occurred in 24 patients (18.0%). Patients with cervix failures had a lower D100, D90, and Dmean than those who did not experience failure in the cervix. The respective doses to the GTV were 41, 58, and 136 Gy for failures compared with 67, 99, and 236 Gy for those who did not experience failure (P<.001). Probit analysis estimated the minimum D100, D90, and Dmean doses required for ≥90% local control to be 69, 98, and 260 Gy (P<.001). Conclusions: Total dose delivered to the GTV from combined MRI-guided HDR and PET/CT-guided IMRT is highly correlated with local tumor control. The findings can be directly applied in the clinic for dose adaptation to maximize local control.

  2. Braquiterapia endoluminal HDR no tratamento de tumores primários ou recidivas na árvore traqueobrônquica Endoluminal high dose rate brachytherapy in the treatment of primary and recurrent bronchogenic tree malignancies

    Directory of Open Access Journals (Sweden)

    Maria Fortunato

    2009-03-01

    primários sintomáticos apresentam uma boa tolerância e um alívio sintomático associado a uma boa qualidade de vida. Apesar da reduzida amostra, os resultados demonstram as eventuais vantagens da BTE de HDR no tratamento paliativo/ curativo destes doentesIntroduction: Locally advanced tumours as the initial form of presentation of tumours in the bronchial tree are not a rare event. Bronchogenic recurrence is frequent in the natural history of some tumours. The choice of therapeutic options from the raft available depends on such variables as initial therapy, place of recurrence, symptoms and patient’s physical status. Aim: To demonstrate the advantages of endoluminal brachytherapy (EBT with high dose rate (HDR in primary and recurrent tumour of the bronchial tree. Material and methods: A retrospective study of seven patients (pts with primary tumours of the colon, trachea and lung. Tracheobronchial recurrence (trachea, two pts, bronchus, five pts occurred between March 2003 and September 2004. Patients under-went EBT with HDR for primary or recurrent therapy in association with external radiotherapy, laser therapy and chemotherapy with palliative or curative intention. EBT with HDR doses of 5 to 7 Gy in 2 to 4 fractions at 1 cm from the source axis were given. Treatment included endoluminal application of Ir192 with a French 6 catheter. Results: There was symptomatic relief related to reduction of tumour in six of the seven patients treated. In one of the six patients studied, there was progression of the local disease between the second and third fractions of the treatment (obstruction of the trachea. In a mean follow up of 17 (2-40 months between EBT and this study, three patients are alive, one has no evidence of disease while two have had bronchial recurrence, four patients have died, one after massive haemoptysis and three due to disease progression. Discussion and conclusions: Patients undergoing brachytherapy for symptomatic primary tumours or

  3. 3-D conformal treatment of prostate cancer to 74 Gy vs. high-dose-rate brachytherapy boost: A cross-sectional quality-of-life survey

    Energy Technology Data Exchange (ETDEWEB)

    Vordermark, Dirk [Univ. of Wuerzburg (DE). Dept. of Radiation Oncology] (and others)

    2006-09-15

    The effects of two modalities of dose-escalated radiotherapy on health-related quality of life (HRQOL) were compared. Forty-one consecutive patients were treated with a 3-D conformal (3-DC) boost to 74 Gy, and 43 with high-dose rate (HDR) brachytherapy boost (2x9 Gy), following 3-D conformal treatment to 46 Gy. Median age was 70 years in both groups, median initial PSA was 7.9 {mu}g/l in 3-DC boost patients and 8.1 {mu}g/l in HDR boost patients. Stage was 7 in 52% and 47%, respectively. HRQOL was assessed cross-sectionally using EORTC QLQ-C30 and organ-specific PR25 modules 3-32 (median 19) and 4-25 (median 14) months after treatment, respectively. Questionnaires were completed by 93% and 97% of patients, respectively. Diarrhea and insomnia scores were significantly increased in both groups. In the PR25 module, scores of 3-DC boost and HDR boost patients for urinary, bowel and treatment-related symptoms were similar. Among responders, 34% of 3-DC boost patients and 86% of HDR boost patients had severe erectile problems. Dose escalation in prostate cancer by either 3-DC boost to 74 Gy or HDR brachytherapy boost appears to result in similar HRQOL profiles.

  4. A rectum shield for the circular applicator system of a selectron unit (HDR and LDR afterloading)

    International Nuclear Information System (INIS)

    Hetzel, H.; McCoy, M.; Kamleitner, H.; Frommhold, H.

    1987-01-01

    In order to decrease the morbidity rate after combined radiotherapy of the cervix carcinoma, a tungsten shield 3 and 5 mm thick for the rectum has been developed by the authors which is applied with the ring and pin applicator of the selectron unit (LDR and HDR afterloading). The isodose curves were measured in a plexiglas phantom, and the radiation dose at the reference points was determined by means of a ionization dosemeter. The phantom measurements were performed with the same arrangement of sources as applied in radiotherapy. The measurements showed a dose reduction at point Rmax of 33% (HDR) and 44% (LDR) with the tungsten shield 5 mm thick. (orig.) [de

  5. Dosimetric intercomparison of permanent Ho-166 seed's implants and HDR Ir-192 brachytherapy in breast cancer.

    Science.gov (United States)

    de Campos, Tarcisio Passos Ribeiro; Nogueira, Luciana Batista; Trindade, Bruno; Cuperschmid, Ethel Mizrahy

    2016-01-01

    To provide a comparative dosimetric analysis of permanent implants of Ho(166)-seeds and temporary HDR Ir(192)-brachytherapy through computational simulation. Brachytherapy with Ir(192)-HDR or LDR based on temporary wires or permanent radioactive seed implants can be used as dose reinforcement for breast radiation therapy. Permanent breast implants have not been a practical clinical routine; although, I(125) and Pd(103)-seeds have already been reported. Biodegradable Ho(166)-ceramic-seeds have been addressed recently. Simulations of implants of nine Ho(166)-seeds and equivalent with HDR Ir(192)-brachytherapy were elaborated in MCNP5, shaped in a computational multivoxel simulator which reproduced a female thorax phantom. Spatial dose rate distributions and dose-volume histograms were generated. Protocol's analysis involving exposure time, seed's activities and dose were performed. Permanent Ho(166)-seed implants presented a maximum dose rate per unit of contained activity (MDR) of 1.1601 μGy h(-1) Bq(-1); and, a normalized MDR in standard points (8 mm, equidistant to 03-seeds - SP1, 10 mm - SP2) of 1.0% (SP1) and 0.5% (SP2), respectively. Ir(192)-brachytherapy presented MDR of 4.3945 × 10(-3) μGy h(-1) Bq(-1); and, 30% (SP1), and 20% (SP2). Therefore, seed's implant activities of 333 MBq (Ho(166)) and 259 GBq (Ir(192)) produced prescribed doses of 58 Gy (SP1; 5d) and 56 Gy (SP1, 5 fractions, 6 min), respectively. Breast Ho(166)-implants of 37-111 MBq are attractive due to the high dose rate near 6-10 mm from seeds, equivalent to Ir(192)-brachytherapy of 259 GBq (3 fractions, 6 min) providing similar dose in standard points at a week; however, with spatial dose distribution better confined. The seed positioning can be adjusted for controlling the breast tumor, in stages I and II, in flat and deep tumors, without any breast volumetric limitation.

  6. Computed tomography-guided interstitial high dose rate brachytherapy for centrally located liver tumours: a single institution study

    Energy Technology Data Exchange (ETDEWEB)

    Tselis, Nikolaos; Chatzikonstantinou, Georgios; Zamboglou, Nikolaos [Klinikum Offenbach, Department of Radiation Oncology, Offenbach am Main (Germany); Kolotas, Christos [Hirslanden Medical Center, Institute for Radiotherapy, Aarau (Switzerland); Milickovic, Natasa; Baltas, Dimos [Klinikum Offenbach, Department of Medical Physics and Engineering, Offenbach am Main (Germany)

    2013-08-15

    To evaluate the clinical outcome of computed tomography (CT)-guided interstitial (IRT) high-dose-rate (HDR) brachytherapy (BRT) in the treatment of unresectable primary and secondary liver malignancies. This report updates and expands our previously described experience with this treatment technique. Forty-one patients with 50 tumours adjacent to the liver hilum and bile duct bifurcation were treated in 59 interventions of CT-guided IRT HDR BRT. The tumours were larger than 4 cm with a median volume of 84 cm{sup 3} (38-1,348 cm{sup 3}). The IRT HDR BRT delivered a median total physical dose of 20.0 Gy (7.0-32.0 Gy) in twice daily fractions of median 7.0 Gy (4.0-10.0 Gy) in 19 patients and in once daily fractions of median 8.0 Gy (7.0-14.0 Gy) in 22 patients. With a median follow-up of 12.4 months, the local control for metastatic hepatic tumours was 89 %, 73 % and 63 % at 6, 12 and 18 months respectively. The local control for primary hepatic tumours was 90 %, 81 % and 50 % at 6, 12 and 18 months respectively. Severe side effects occurred in 5.0 % of interventions with no treatment-related deaths. CT-guided IRT HDR BRT is a promising procedure for the radiation treatment of centrally located liver malignancies. (orig.)

  7. Rectal doses during LDR and HDR intracavitary brachytherapy of gyneacological malignancies: comparison of direct measurement with that of calculated from radiograph

    International Nuclear Information System (INIS)

    Chougule, Arun; Agarwal, D.P.

    2001-01-01

    In the present study rectal doses using CaSO 4 :Dy powder has been measured in 14 cases of cancer cervix treated by LDR brachytherapy and 20 cases of cancer cervix treated by HDR brachytherapy. The maximum rectal dose in LDR varied from 1073-1856 cGy for point A dose of 3000 cGy. The maximum rectal dose was found to be at 6-8 cm from the anal verge. The results of the calculation and actual measurements has been compared

  8. Non-uniform dwell times in line source high dose rate brachytherapy: physical and radiobiological considerations

    International Nuclear Information System (INIS)

    Jones, B.; Tan, L.T.; Freestone, G.; Bleasdale, C.; Myint, S.; Littler, J.

    1994-01-01

    The ability to vary source dwell times in high dose rate (HDR) brachytherapy allows for the use of non-uniform dwell times along a line source. This may have advantages in the radical treatment of tumours depending on individual tumour geometry. This study investigates the potential improvements in local tumour control relative to adjacent normal tissue isoeffects when intratumour source dwell times are increased along the central portion of a line source (technique A) in radiotherapy schedules which include a relatively small component of HDR brachytherapy. Such a technique is predicted to increase the local control for tumours of diameters ranging between 2 cm and 4 cm by up to 11% compared with a technique in which there are uniform dwell times along the line source (technique B). There is no difference in the local control rates for the two techniques when used to treat smaller tumours. Normal tissue doses are also modified by the technique used. Technique A produces higher normal tissue doses at points perpendicular to the centre of the line source and lower dose at points nearer the ends of the line source if the prescription point is not in the central plane of the line source. Alternatively, if the dose is prescribed at a point in the central plane of the line source, the dose at all the normal tissue points are lower when technique A is used. (author)

  9. Neoadjuvant androgen deprivation and long-term results for patients with intermediate- and high-risk prostate cancer treated with high-dose rate brachytherapy and external beam radiotherapy

    International Nuclear Information System (INIS)

    Pellizzon, Antonio Cassio Assis; Fogaroli, Ricardo Cesar; Silva, Maria Leticia Gobo; Castro, Douglas Guedes; Maia, Maria Conte

    2010-01-01

    Purpose: to evaluate the influence of neoadjuvant androgen deprivation (NAAD) and report the long term biochemical control rates according to the Phoenix Consensus Conference, and prognostic factors of intermediate- (IR) and high-risk (HR) prostate cancer treated with external beam radiotherapy and high-dose-rate brachytherapy (HDR-BT). Methods and materials: between March, 1997 and June, 2005, 184 patients considered IR or HR were treated with localized radiotherapy and HDR-BT at the Department of Radiation Oncology, Hospital A.C. Camargo, Sao Paulo, Brazil. Patient's age, Gleason score, clinical stage, initial PSA value, risk group for biochemical failure, presence of NAAD, doses of radiotherapy and HDR-BT were evaluated. Results: median age and follow-up were 70 years old (range, 47-83) and 74.5 months (range, 24-123 months), respectively. Patients considered IR were 91 (49.4%) and HR 93 (50.6%). Ninety-nine (53.8%) patients had no NAAD. The overall survival at 5 years was 93.6%. The 5-year actuarial biochemical control rates for all patients, IR and HR were 83.4%, 86.2% and 78.8%, respectively, p0.076. On univariate analysis the prognostic factors related to better biochemical control were Gleason score 45 Gy (p= 0.011) and HDR-BT dose > 20 Gy (p< 0.001). On multivariate analysis no statistical significant predictive factor related to biochemical control was found. Conclusions: the role of NAAD for IR and HR prostate cancer is still to be defined. HDR-BT combined to external radiotherapy is a successful form of treatment for these patients, with our results comparable to published data. (author)

  10. Second salvage high-dose-rate brachytherapy for radiorecurrent prostate cancer

    Directory of Open Access Journals (Sweden)

    Metha Maenhout

    2017-04-01

    Full Text Available Purpose : Salvage treatments for localized radiorecurrent prostate cancer can be performed safely when a focal and image guided approach is used. Due to the low toxicity, the opportunity exists to investigate a second salvage treatment when a second locally recurrent prostate cancer occurs. Here, we describe a second salvage treatment procedure of 4 patients. Material and methods : Four patients with a pathologically proven second local recurrence were treated in an outpatient magnetic resonance imaging (MRI-guided setting with a single fraction of 19 Gy focal high-dose-rate brachytherapy (HDR-BT. Delineation was performed using choline-PET-CT or a 68Ga-PSMA PET in combination with multiparametric 3 Tesla MRI in all four patients. Toxicity was measured using common toxicity criteria for adverse events (CTCAE version 4.0. Results : With a median follow-up of 12 months (range, 6-15, there were 2 patients with biochemical recurrence as defined by the Phoenix-definition. There were no patients with grade 3 or more toxicity. In all second salvage HDR-BT treatments, the constraints for rectum, bladder, and urethra were met. Median treatment volume (GTV was 4.8 cc (range, 1.9-6.6 cc. A median of 8 catheters (range, 6-9 were used, and the median dose to the treatment volume (GTV was a D95: 19.3 Gy (SD 15.5-19.4 Gy. Conclusions : Second focal salvage MRI-guided HDR-BT for a select group of patients with a second locally recurrent prostate cancer is feasible. There was no grade 3 or more acute toxicity for these four patients.

  11. High dose-rate brachytherapy source localization: positional resolution using a diamond detector

    International Nuclear Information System (INIS)

    Nakano, T; Suchowerska, N; Bilek, M M; McKenzie, D R; Ng, N; Kron, T

    2003-01-01

    A potential real-time source position verification process for high dose-rate (HDR) brachytherapy treatment is described. This process is intended to provide immediate confirmation that a treatment is proceeding according to plan, so that corrective action can be taken if necessary. We show that three dosimeters are in principle sufficient and demonstrate the feasibility of the process using a diamond detector and an Ir-192 source. An error analysis including all identified sources of error shows that this detector is capable of locating the distance to the source to within 2 mm for distances up to 12 cm. This positional accuracy is less than the diameter of typical HDR catheters indicating that a diamond detector can be used to accurately determine the distance to the source. The uncertainty in the distance is found to increase with distance

  12. An Internet-ready database for prospective randomized clinical trials of high-dose-rate brachytherapy for adenocarcinoma of the prostate

    International Nuclear Information System (INIS)

    Devlin, Phillip M.; Brus, Christina R.; Kazakin, Julia; Mitchell, Ronald B.; Demanes, D. Jeffrey; Edmundson, Gregory; Gribble, Michael; Gustafson, Gary S.; Kelly, Douglas A.; Linares, Luis A.; Martinez, Alvaro A.; Mate, Timothy P.; Nag, Subir; Perez, Carlos A.; Rao, Jaynath G.; Rodriguez, Rodney R.; Shasha, Daniel; Tripuraneni, Prabhakar

    2002-01-01

    Purpose: To demonstrate a new interactive Internet-ready database for prospective clinical trials in high-dose-rate (HDR) brachytherapy for prostate cancer. Methods and Materials: An Internet-ready database was created that allows common data acquisition and statistical analysis. Patient anonymity and confidentiality are preserved. These data forms include all common elements found from a survey of the databases. The forms allow the user to view patient data in a view-only or edit mode. Eight linked forms document patient data before and after receiving HDR therapy. The pretreatment forms are divided into four categories: staging, comorbid diseases, external beam radiotherapy data, and signs and symptoms. The posttreatment forms separate data by HDR implant information, HDR medications, posttreatment signs and symptoms, and follow-up data. The forms were tested for clinical usefulness. Conclusion: This Internet-based database enables the user to record and later analyze all relevant medical data and may become a reliable instrument for the follow-up of patients and evaluation of treatment results

  13. Postoperative high dose rate vaginal apex brachytherapy in stage I endometrial adenocarcinoma

    International Nuclear Information System (INIS)

    Turner, B.; Knisely, J.; Kacinski, B.; Roberts, K.; Peschel, R.; Gumbs, A.; Rutherford, T.; Edraki, B.; Schwartz, P.; Chambers, J.; Kohorn, E.; Wilson, L.

    1996-01-01

    Introduction: Patients with stage I endometrial adenocarcinoma have traditionally been treated with TAH/BSO and radiation. The reported incidence of local recurrence in surgically treated patients with Figo stage IA or IB endometrial adenocarcinoma is 4-10% at 2 years. Combined surgery and radiation has resulted in a reduction of recurrence to 2-6%. We report the presentation, local and distant control, salvage rate, survival and complications for patients undergoing high dose rate (HDR) vaginal apex brachytherapy following surgery. Methods: Between 1985 and 1994 a total of 302 patients with Figo stage I endometrial carcinoma were treated with HDR Ir-192 vaginal apex brachytherapy to a total dose of 21 Gy in 3 fractions at 0.5 cm from the vaginal mucosa. The pathologic stage by treatment group was IA: 31%, IB: 68%, and IC: 1%. The histologic grade of the patient's tumors was grade 1: 69%, 2: 27%, and 3: 4%. The median time from surgery to radiation was 33 days (range 14-66 days). The median follow-up for 300 patients with stage IA (91 patients), IB (205 patients), and IC (4 patients) was respectively 36, 34 and 40 months (2 patients lost to follow-up prior to 6 months). Results: Patients presented with vaginal bleeding (94%) or abnormal pap smear (6%) at a median age for stage IA and IB, of 55 and 64 years, respectively. The crude overall survival of the patient population at 2 years is 95%. Median overall time to failure is 19.5 months (range 10-36 months). The overall failure rate was 2.7% (8 patients), local failure only 1.0% (38% of failed group), distant failure only 0.3% (12% of failed group) and combined local/distant failure 1.3% (50% of failed group). The local failure rate for pathologic stage IA patients was 1.0% and no distant disease was observed. The local failure rate for pathologic stage IB patients was 3.4% 7/205 and distant failure was 2.4% 5/205. The majority of patients with recurrence had grade 2 histologic changes 5/8. The overall salvage rate

  14. High-dose rate brachytherapy in the treatment of carcinoma of uterine cervix: twenty-year experience with cobalt after-loading system.

    Science.gov (United States)

    Mosalaei, A; Mohammadianpanah, M; Omidvari, S; Ahmadloo, N

    2006-01-01

    This retrospective analysis aims to report results of patients with cancer of uterine cervix treated with external-beam radiotherapy (EBR) and high-dose rate (HDR) brachytherapy, using manual treatment planning. From 1975 to 1995, 237 patients with FIGO stages IIB-IVA and mean age of 54.31 years were treated. EBR dose to the whole pelvis was 50 Gy in 25 fractions. Brachytherapy with HDR after-loading cobalt source (Cathetron) was performed following EBR completion with a dose of 30 Gy in three weekly fractions of 10 Gy to point A. Survival, local control, and genitourinary and gastrointestinal complications were assessed. In a median follow-up of 60.2 months, the 10-year overall and disease-free survival rate was 62.4%. Local recurrence was seen in 12.2% of patients. Distant metastases to the lymph nodes, peritoneum, lung, liver, and bone occurred in 25.3% of patients. Less than 6% of patients experienced severe genitourinary and/or gastrointestinal toxicity that were relieved by surgical intervention. No treatment-related mortality was seen. This series suggests that 50 Gy to the whole pelvis together with three fractions of 10 Gy to point A with HDR brachytherapy is an effective fractionation schedule in the treatment of locally advanced cancer of cervix. To decrease the complications, newer devices and treatment planning may be beneficial.

  15. 3-D conformal HDR brachytherapy as monotherapy for localized prostate cancer. A pilot study

    International Nuclear Information System (INIS)

    Martin, T.; Baltas, D.; Kurek, R.; Roeddiger, S.; Kontova, M.; Anagnostopoulos, G.; Skazikis, G.; Zamboglou, N.; Dannenberg, T.; Buhleier, T.; Tunn, U.

    2004-01-01

    Purpose: pilot study to evaluate feasibility, acute toxicity and conformal quality of three-dimensional (3-D) conformal high-dose-rate (HDR) brachytherapy as monotherapy for localized prostate cancer using intraoperative real-time planning. Patients and methods: between 05/2002 and 05/2003, 52 patients with prostate cancer, prostate-specific antigen (PSA) ≤ 10 ng/ml, Gleason score ≤ 7 and clinical stage ≤ T2a were treated. Median PSA was 6.4 ng/ml and median Gleason score 5. 24/52 patients had stage T1c and 28/52 stage T2a. For transrectal ultrasound-(TRUS-)guided transperineal implantation of flexible plastic needles into the prostate, the real-time HDR planning system SWIFT trademark was used. After implantation, CT-based 3-D postplanning was performed. All patients received one implant for four fractions of HDR brachytherapy in 48 h using a reference dose (D ref ) of 9.5 Gy to a total dose of 38.0 Gy. Dose-volume histograms (DVHs) were analyzed to evaluate the conformal quality of each implant using D 90 , D 10 urethra, and D 10 rectum. Acute toxicity was evaluated using the CTC (common toxicity criteria) scales. Results: median D 90 was 106% of D ref (range: 93-115%), median D 10 urethra 159% of D ref (range: 127-192%), and median D 10 rectum 55% of D ref (range: 35-68%). Median follow-up is currently 8 months. In 2/52 patients acute grade 3 genitourinary toxicity was observed. No gastrointestinal toxicity > grade 1 occurred. Conclusion: 3-D conformal HDR brachytherapy as monotherapy using intraoperative real-time planning is a feasible and highly conformal treatment for localized prostate cancer associated with minimal acute toxicity. Longer follow-up is needed to evaluate late toxicity and biochemical control. (orig.)

  16. Phase II trial of combined surgical resection, high dose rate intraoperative radiation therapy, and external beam radiotherapy for malignant pleural mesothelioma

    International Nuclear Information System (INIS)

    Raben, A.; Rusch, V.; Mychalczak, B.; Ginsberg, R.; Burt, M.; Bains, M.; Francois, Damien; Harrison, L.B.

    1997-01-01

    Purpose: To determine the feasibility of combining extrapleural pneumonectomy (EPP) or pleurectomy / decortication (PD), high dose rate intraoperative radiation therapy (HDR-IORT) and postoperative external beam radiation hemithoracic radiation (EBHRT) to treat malignant pleural mesothelioma (MPM). Materials and Methods: From 3/94 through 9/94, 15 patients (pts) were enrolled on this trial. This included 3 females and 12 males with a median age of 59 (Range: 45-75). Eligibility criteria included biopsy proven MPM, no evidence of T4 or N3 disease by exam/CT/MRI, no evidence of metastatic disease, no previous treatment, and a Karnofsky performance status of ≥ 80%. Pts with pulmonary function tests permitting EPP, underwent EPP and HDR-IORT (N=7). The rest underwent PD/HDR-IORT (N=4). An intraoperative dose of 15 Gy was prescribed to a depth of 5 mm in tissue to the ipsilateral mediastinum, diaphragm, and chest wall. Postoperatively, 54 Gy of EBHRT was prescribed to the hemithorax, surgical scar and surgical drain site. The median surgical procedure time, median IORT time and median overall operating time was 554 minutes, 240 minutes, and 649 minutes respectively. The median dose of EBHRT was 50.4 Gy (Range 50-54 Gy). The median follow-up time is 8 months (Range: 3.5 to 28 months). Four of 15 pts had unresectable disease at the time of surgery and were taken off study. Results are presented in crude and actuarial analysis. Results: A complete resection of all visible gross disease was accomplished in 10 pts. One pt had a single focus of gross residual disease (less than 5 mm in size) left behind in the chest wall. The overall complication rate was 54%. Treatment related mortality occurred in 2 pts (18%) at 1 and 7 months respectively. This was attributed to ARDS in 1 pt (EPP/HDR-IORT) and radiation pneumonitis combined with a tracheoesophageal fistula in 1 pt (PD/HDR-IORT). Of the 6 remaining pts undergoing EPP/HDR-IORT, 2 pts developed a postoperative empyema with

  17. Novel Use of the Contura for High Dose Rate Cranial Brachytherapy

    International Nuclear Information System (INIS)

    Scanderbeg, Daniel J.; Alksne, John F.; Lawson, Joshua D.; Murphy, Kevin T.

    2011-01-01

    A popular choice for treatment of recurrent gliomas was cranial brachytherapy using the GliaSite Radiation Therapy System. However, this device was taken off the market in late 2008, thus leaving a treatment void. This case study presents our experience treating a cranial lesion for the first time using a Contura multilumen, high-dose-rate (HDR) brachytherapy balloon applicator. The patient was a 47-year-old male who was diagnosed with a recurrent right frontal anaplastic oligodendroglioma. Previous radiosurgery made him a good candidate for brachytherapy. An intracavitary HDR balloon brachytherapy device (Contura) was placed in the resection cavity and treated with a single fraction of 20 Gy. The implant, treatment, and removal of the device were all completed without incident. Dosimetry of the device was excellent because the dose conformed very well to the target. V90, V100, V150, and V200 were 98.9%, 95.7%, 27.2, and 8.8 cc, respectively. This patient was treated successfully using the Contura multilumen balloon. Contura was originally designed for deployment in a postlumpectomy breast for treatment by accelerated partial breast irradiation. Being an intracavitary balloon device, its similarity to the GliaSite system makes it a viable replacement candidate. Multiple lumens in the device also make it possible to shape the dose delivered to the target, something not possible before with the GliaSite applicator.

  18. Physics and quality assurance for high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Anderson, Lowell L.

    1995-01-01

    Purpose: To review the physical aspects of high dose rate (HDR) brachytherapy, including commissioning and quality assurance, source calibration and dose distribution measurements, and treatment planning methods. Following the introduction of afterloading in brachytherapy, development efforts to make it 'remote' culminated in 1964 with the near-simultaneous appearance of remote afterloaders in five major medical centers. Four of these machines were 'high dose rate', three employing 60Co and one (the GammaMed) using a single, cable-mounted 192Ir source. Stepping-motor source control was added to the GammaMed in 1974, making it the precursor of modern remote afterloaders, which are now suitable for interstitial as well as intracavitary brachytherapy by virtue of small source-diameter and indexer-accessed multiple channels. Because the 192Ir sources currently used in HDR remote afterloaders are supplied at a nominal air-kerma strength of 11.4 cGy cm2 s-1 (10 Ci), are not collimated in clinical use, and emit a significant fraction (15%) of photons at energies greater than 600 keV, shielding and facility design must be undertaken as carefully and thoroughly as for external beam installations. Licensing requirements of regulatory agencies must be met with respect both to maximum permissible dose limits and to the existence and functionality of safety devices (door interlocks, radiation monitors, etc.). Commissioning and quality assurance procedures that must be documented for HDR remote afterloading relate to (1) machine, applicator, guide-tube, and facility functionality checks, (2) source calibration, (3) emergency response readiness, (4) planning software evaluation, and (5) independent checks of clinical dose calculations. Source calibration checks must be performed locally, either by in-air measurement of air kerma strength or with a well ionization chamber calibrated (by an accredited standards laboratory) against an in-air measurement of air kerma strength for the

  19. Management of a HDR brachytherapy system in the Hospital Juarez of Mexico

    International Nuclear Information System (INIS)

    Serrano F, A.G.; Ramirez R, G.; Gil G, R.; Azorin N, J.; Rivera M, T.

    2007-01-01

    Full text: In the Hospital Juarez of Mexico, it is carried out a project to implement a Brachytherapy system with high dose rate (HDR) through a Management quality program. In our work center this treatment modality in patients with cervicouterine cancer is used (CaCu), and constantly it is necessary to carry out improvements in the procedures, with the purpose of optimizing them and in consequence to complete the principles of the Radiological Protection, guaranteeing this way, an attention with the quality and safety, such that allow to diminish the risks to the patients and to assure that the received dose in critical organs it finds inside the permitted therapeutic limits, without commit the radiosensitive response of healthy organs. In this work an analysis of the implementation of this system is presented, detailing the procedures so much in the technological infrastructure like human and indicating the necessary technical and operative requirements to reach an adequate practice in HDR brachytherapy. (Author)

  20. Implementation of the technique of partial irradiation accelerated the breast with high doses (HDR) brachytherapy

    International Nuclear Information System (INIS)

    Molina Lopez, M. Y.; Pardo Perez, E.; Castro Novais, J.; Martinez Ortega, J.; Ruiz Maqueda, S.; Cerro Penalver, E. del

    2013-01-01

    The objective of this work is presents procedure carried out in our Centre for the implementation of the accelerated partial breast irradiation (APBI, accelerated partial-breast irradiation) with high-rate brachytherapy (HDR), using plastic tubes as applicators. Carried out measures, the evaluation of the dosimetric parameters analyzing and presenting the results. (Author)

  1. Nationwide, Multicenter, Retrospective Study on High-Dose-Rate Brachytherapy as Monotherapy for Prostate Cancer

    International Nuclear Information System (INIS)

    Yoshioka, Yasuo; Kotsuma, Tadayuki; Komiya, Akira; Kariya, Shinji; Konishi, Koji; Nonomura, Norio; Ogawa, Kazuhiko; Tanaka, Eiichi; Nishimura, Kensaku; Fujiuchi, Yasuyoshi; Kitamura, Hiroshi; Yamagami, Takuji; Yamasaki, Ichiro; Nishimura, Kazuo; Teshima, Teruki; Nakamura, Katsumasa; Itami, Jun

    2017-01-01

    Purpose: To present, analyze, and discuss results of a nationwide, multicenter, retrospective study on high-dose-rate brachytherapy (HDR-BT) as monotherapy for low-, intermediate-, and high-risk prostate cancer. Methods and Materials: From 1995 through 2013, 524 patients, 73 (14%) with low-risk, 207 (40%) with intermediate-risk, and 244 (47%) with high-risk prostate cancer, were treated with HDR-BT as monotherapy at 5 institutions in Japan. Dose fractionations were 27 Gy/2 fractions for 69 patients (13%), 45.5 Gy/7 fractions for 168 (32%), 49 Gy/7 fractions for 149 (28%), 54 Gy/9 fractions for 130 (25%), and others for 8 (2%). Of these patients, 156 (30%) did not receive androgen deprivation therapy, and 202 patients (39%) did receive androgen deprivation therapy 3 years. Median follow-up time was 5.9 years (range, 0.4-18.1 years), with a minimum of 2 years for surviving patients. Results: After 5 years, respective actuarial rates of no biochemical evidence of disease, overall survival, cause-specific survival, and metastasis-free survival for all patients were 92%, 97%, 99%, and 94%. For low/intermediate/high-risk patients, the 5-year no biochemical evidence of disease rates were 95%/94%/89%, the 5-year overall survival rates were 98%/98%/94%, the 5-year cause-specific survival rates were 98%/100%/98%, and the 5-year metastasis-free survival rates were 98%/95%/90%, respectively. The cumulative incidence of late grade 2 to 3 genitourinary toxicity at 5 years was 19%, and that of late grade 3 was 1%. The corresponding incidences of gastrointestinal toxicity were 3% and 0% (0.2%). No grade 4 or 5 of either type of toxicity was detected. Conclusions: The findings of this nationwide, multicenter, retrospective study demonstrate that HDR-BT as monotherapy was safe and effective for all patients with low-, intermediate-, and high-risk prostate cancer.

  2. Nationwide, Multicenter, Retrospective Study on High-Dose-Rate Brachytherapy as Monotherapy for Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Yoshioka, Yasuo, E-mail: yoshioka@radonc.med.osaka-u.ac.jp [Department of Radiation Oncology, Osaka University Graduate School of Medicine, Osaka (Japan); Kotsuma, Tadayuki [Department of Radiation Oncology, Osaka National Hospital, Osaka (Japan); Komiya, Akira [Department of Urology, Graduate School of Medicine and Pharmaceutical Sciences for Research, University of Toyama, Toyama (Japan); Department of Urology, Chiba University Hospital, Chiba (Japan); Kariya, Shinji [Department of Radiology, Kochi Medical School, Kochi (Japan); Konishi, Koji [Department of Radiation Oncology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka (Japan); Nonomura, Norio [Department of Urology, Osaka University Graduate School of Medicine, Osaka (Japan); Ogawa, Kazuhiko [Department of Radiation Oncology, Osaka University Graduate School of Medicine, Osaka (Japan); Tanaka, Eiichi [Department of Radiation Oncology, Osaka National Hospital, Osaka (Japan); Nishimura, Kensaku [Department of Urology, Osaka National Hospital, Osaka (Japan); Fujiuchi, Yasuyoshi; Kitamura, Hiroshi [Department of Urology, Graduate School of Medicine and Pharmaceutical Sciences for Research, University of Toyama, Toyama (Japan); Yamagami, Takuji [Department of Radiology, Kochi Medical School, Kochi (Japan); Yamasaki, Ichiro [Department of Urology, Kochi Medical School, Kochi (Japan); Nishimura, Kazuo [Department of Urology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka (Japan); Teshima, Teruki [Department of Radiation Oncology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka (Japan); Nakamura, Katsumasa [Department of Radiation Oncology, Hamamatsu University School of Medicine, Shizuoka (Japan); Itami, Jun [Department of Radiation Oncology, National Cancer Center Hospital, Tokyo (Japan)

    2017-04-01

    Purpose: To present, analyze, and discuss results of a nationwide, multicenter, retrospective study on high-dose-rate brachytherapy (HDR-BT) as monotherapy for low-, intermediate-, and high-risk prostate cancer. Methods and Materials: From 1995 through 2013, 524 patients, 73 (14%) with low-risk, 207 (40%) with intermediate-risk, and 244 (47%) with high-risk prostate cancer, were treated with HDR-BT as monotherapy at 5 institutions in Japan. Dose fractionations were 27 Gy/2 fractions for 69 patients (13%), 45.5 Gy/7 fractions for 168 (32%), 49 Gy/7 fractions for 149 (28%), 54 Gy/9 fractions for 130 (25%), and others for 8 (2%). Of these patients, 156 (30%) did not receive androgen deprivation therapy, and 202 patients (39%) did receive androgen deprivation therapy <1 year, 112 (21%) for 1-3 years, and 54 (10%) for >3 years. Median follow-up time was 5.9 years (range, 0.4-18.1 years), with a minimum of 2 years for surviving patients. Results: After 5 years, respective actuarial rates of no biochemical evidence of disease, overall survival, cause-specific survival, and metastasis-free survival for all patients were 92%, 97%, 99%, and 94%. For low/intermediate/high-risk patients, the 5-year no biochemical evidence of disease rates were 95%/94%/89%, the 5-year overall survival rates were 98%/98%/94%, the 5-year cause-specific survival rates were 98%/100%/98%, and the 5-year metastasis-free survival rates were 98%/95%/90%, respectively. The cumulative incidence of late grade 2 to 3 genitourinary toxicity at 5 years was 19%, and that of late grade 3 was 1%. The corresponding incidences of gastrointestinal toxicity were 3% and 0% (0.2%). No grade 4 or 5 of either type of toxicity was detected. Conclusions: The findings of this nationwide, multicenter, retrospective study demonstrate that HDR-BT as monotherapy was safe and effective for all patients with low-, intermediate-, and high-risk prostate cancer.

  3. SU-F-T-14: Dosimetric Impacts of Various Uncertainties in Cervical Cancer HDR Brachytherapy: Are Conventional Point Doses Good Surrogates for 3D Dosimetry?

    Energy Technology Data Exchange (ETDEWEB)

    Liang, X; Li, Z [University of Florida Health Proton Therapy Institute, Jacksonville, FL (United States); Zheng, D [University of Nebraska Medical Center, Omaha, NE (United States); Zhang, X; Narayanasamy, G; Morrill, S; Penagaricano, J; Paudel, N [University of Arkansas for Medical Sciences, Little Rock, AR (United States)

    2016-06-15

    Purpose: In the context of evaluating dosimetric impacts of a variety of uncertainties involved in HDR Tandem-and-Ovoid treatment, to study the correlations between conventional point doses and 3D volumetric doses. Methods: For 5 cervical cancer patients treated with HDR T&O, 150 plans were retrospectively created to study dosimetric impacts of the following uncertainties: (1) inter-fractional applicator displacement between two treatment fractions within a single insertion by applying Fraction#1 plan to Fraction#2 CT; (2) positional dwell error simulated from −5mm to 5mm in 1mm steps; (3) simulated temporal dwell error of 0.05s, 0.1s, 0.5s, and 1s. The original plans were based on point dose prescription, from which the volume covered by the prescription dose was generated as the pseudo target volume to study the 3D target dose effect. OARs were contoured. The point and volumetric dose errors were calculated by taking the differences between original and simulated plans. The correlations between the point and volumetric dose errors were analyzed. Results: For the most clinically relevant positional dwell uncertainty of 1mm, temporal uncertainty of 0.05s, and inter-fractional applicator displacement within the same insertion, the mean target D90 and V100 deviation were within 1%. Among these uncertainties, the applicator displacement showed the largest potential target coverage impact (2.6% on D90) as well as the OAR dose impact (2.5% and 3.4% on bladder D2cc and rectum D2cc). The Spearman correlation analysis shows a correlation coefficient of 0.43 with a p-value of 0.11 between target D90 coverage and H point dose. Conclusion: With the most clinically relevant positional and temporal dwell uncertainties and patient interfractional applicator displacement within the same insertion, the dose error is within clinical acceptable range. The lack of correlation between H point and 3D volumetric dose errors is a motivator for the use of 3D treatment planning in

  4. WE-DE-201-02: A Statistical Analysis Tool for Plan Quality Verification in HDR Brachytherapy Forward Planning for Cervix Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Ma, R; Zhu, X; Li, S; Zheng, D; Lei, Y; Wang, S; Verma, V; Bennion, N; Wahl, A; Zhou, S [University of Nebraska Medical Center, Omaha, NE (United States)

    2016-06-15

    Purpose: High Dose Rate (HDR) brachytherapy forward planning is principally an iterative process; hence, plan quality is affected by planners’ experiences and limited planning time. Thus, this may lead to sporadic errors and inconsistencies in planning. A statistical tool based on previous approved clinical treatment plans would help to maintain the consistency of planning quality and improve the efficiency of second checking. Methods: An independent dose calculation tool was developed from commercial software. Thirty-three previously approved cervical HDR plans with the same prescription dose (550cGy), applicator type, and treatment protocol were examined, and ICRU defined reference point doses (bladder, vaginal mucosa, rectum, and points A/B) along with dwell times were collected. Dose calculation tool then calculated appropriate range with a 95% confidence interval for each parameter obtained, which would be used as the benchmark for evaluation of those parameters in future HDR treatment plans. Model quality was verified using five randomly selected approved plans from the same dataset. Results: Dose variations appears to be larger at the reference point of bladder and mucosa as compared with rectum. Most reference point doses from verification plans fell between the predicted range, except the doses of two points of rectum and two points of reference position A (owing to rectal anatomical variations & clinical adjustment in prescription points, respectively). Similar results were obtained for tandem and ring dwell times despite relatively larger uncertainties. Conclusion: This statistical tool provides an insight into clinically acceptable range of cervical HDR plans, which could be useful in plan checking and identifying potential planning errors, thus improving the consistency of plan quality.

  5. High-Dose-Rate Brachytherapy and External-Beam Radiotherapy for Hormone-Naïve Low- and Intermediate-Risk Prostate Cancer: A 7-Year Experience

    International Nuclear Information System (INIS)

    Aluwini, Shafak; Rooij, Peter H. van; Kirkels, Wim J.; Jansen, Peter P.; Praag, John O.; Bangma, Chris H.; Kolkman-Deurloo, Inger-Karine K.

    2012-01-01

    Purpose: To report clinical outcomes and early and late complications in 264 hormone-naïve patients with low- and intermediate-risk prostate cancer treated with high-dose-rate brachytherapy (HDR-BT) in combination with external-beam radiotherapy (EBRT). Methods and Materials: Between February 2000 and July 2007, 264 patients underwent HDR-BT in combination with EBRT as a treatment for their low- to intermediate-risk prostate cancer. The HDR-BT was performed using ultrasound-based implantation. The total HDR-BT dose was 18 Gy in 3 fractions within 24 h, with a 6-h minimum interval. The EBRT started 2 weeks after HDR-BT and was delivered in 25 fractions of 1.8 Gy to 45 Gy within 5 weeks. Results: After a mean follow-up of 74.5 months, 4 patients (1.5%) showed prostate-specific antigen progression according to the American Society for Radiation Oncology definition and 8 patients (3%) according to the Phoenix definition. A biopsy-proven local recurrence was registered in 1 patient (0.4%), and clinical progression (bone metastases) was documented in 2 patients (0.7%). Seven-year actuarial freedom from biochemical failure was 97%, and 7-year disease-specific survival and overall survival were 100% and 91%, respectively. Toxicities were comparable to other series. Conclusions: Treatment with interstitial HDR-BT plus EBRT shows a low incidence of late complications and a favorable oncologic outcome after 7 years follow-up.

  6. SU-D-19A-07: Dosimetric Comparison of HDR Plesiotherapy and Electron Beam Therapy for Superficial Lesions

    Energy Technology Data Exchange (ETDEWEB)

    Mitchell, A; Jacob, D; Andreou, K; Raben, A; Chen, H; Koprowski, C; Mourtada, F [Christiana Care Hospital, Newark, DE (United States)

    2014-06-01

    Purpose: Large superficial (skin, soft tissue sarcoma) lesions located on curved areas are hard to treat with electrons. The Freiburg Flap (Nucletron, Netherlands) is a flexible mesh style surface which can be easily shaped to fit curved surfaces for reproducible HDR fraction delivery. To understand the fundamental dosimetric differences, a dosimetric comparison was made between HDR plesiotherapy (Freiburg applicator for lesions over 4cm) and external electron beam radiotherapy over cases with varying target curvature (both stylized and clinical cases). Methods: Four stylized cases with variable complexity were created using artificial DICOM axial CT slices and RT structures (a square and three curved structures on a 4.5cm radius cylinder). They were planned using Oncentra v4.3 and exported to Pinnacle v9.6 for electrons planning. The HDR source dwell positions were optimized for the best coverage of the targets using graphical optimization. Electron treatment plans were created in Pinnacle using the same CT and RT structures of three HDR cases with surface lesions previously treated with the Freiburg flap. The En face electron plans used 6-12 MeV electrons and 0.5–1 cm bolus was added to increase surface dose. The electron plans were prescribed to an isodose line to conform to the target. Results: For all lesions, the average target dose coverage was similar (D90ave of 100% for HDR vs 101% for electrons). For lesions with high curvature, the HDR coverage was better (D90 102% vs D90 97% for electron). For all cases, adjacent structures high dose region was lower for HDR than electrons (D1cc 100% for HDR vs D1cc 111% for electrons). Conclusion: HDR plesiotherapy offers excellent target conformity for superficial targets similar to electrons. However, for lesions with complex curved surfaces, HDR has the advantage to achieve better dose distributions using graphical optimization to spare adjacent normal tissue while maximizing target coverage.

  7. Needle displacement during HDR brachytherapy in the treatment of prostate cancer

    International Nuclear Information System (INIS)

    Damore, Steven J.; Syed, A.M. Nisar; Puthawala, Ajmel A.; Sharma, Anil

    2000-01-01

    Purpose: We used clinical patient data to examine implant displacement between high dose rate (HDR) brachytherapy fractions for prostate cancer to determine its impact on treatment delivery. Materials and Methods: We analyzed the verification films taken prior to each fraction for 96 consecutive patients treated with HDR brachytherapy boosts as part of their radiation therapy for definitive treatment of organ-confined prostate cancer at our institution. Patients were treated with 18-24 Gy in 4 fractions of HDR delivered in 40 hours followed by 36-39.6 Gy external beam radiation to the prostate. We determined the mean and maximum displacement distances of marker seeds placed in the prostate and of the implanted needles between HDR fractions. Results: Mean and maximum displacement distances between fractions were documented up to 7.6 mm and 28.5 mm, respectively, for the implant needles and 3.6 mm and 11.4 mm, respectively, for the gold marker seeds. All displacement of implant needles occurred in the caudal direction. At least 1 cm caudal displacement of needles occurred prior to 15.5% all fractions. Manual adjustment of needles was required prior to 15% of fractions, and adjustment of the CLP only was required in 24%. Most of the displacement for both the marker seeds and needles occurred between the first and second fractions. Conclusions: There is significant caudal displacement of interstitial implant needles between HDR fractions in our prostate cancer patients. Obtaining verification films and making adjustments in the treatment volume prior to each fraction is necessary to avoid significant inaccuracies in treatment delivery

  8. SU-F-T-24: Impact of Source Position and Dose Distribution Due to Curvature of HDR Transfer Tubes

    Energy Technology Data Exchange (ETDEWEB)

    Khan, A; Yue, N [Rutgers University, New Brunswick, NJ (United States)

    2016-06-15

    Purpose: Brachytherapy is a highly targeted from of radiotherapy. While this may lead to ideal dose distributions on the treatment planning system, a small error in source location can lead to change in the dose distribution. The purpose of this study is to quantify the impact on source position error due to curvature of the transfer tubes and the impact this may have on the dose distribution. Methods: Since the source travels along the midline of the tube, an estimate of the positioning error for various angles of curvature was determined using geometric properties of the tube. Based on the range of values a specific shift was chosen to alter the treatment plans for a number of cervical cancer patients who had undergone HDR brachytherapy boost using tandem and ovoids. Impact of dose to target and organs at risk were determined and checked against guidelines outlined by radiation oncologist. Results: The estimate of the positioning error was 2mm short of the expected position (the curved tube can only cause the source to not reach as far as with a flat tube). Quantitative impact on the dose distribution is still in the process of being analyzed. Conclusion: The accepted positioning tolerance for the source position of a HDR brachytherapy unit is plus or minus 1mm. If there is an additional 2mm discrepancy due to tube curvature, this can result in a source being 1mm to 3mm short of the expected location. While we do always attempt to keep the tubes straight, in some cases such as with tandem and ovoids, the tandem connector does not extend as far out from the patient so the ovoid tubes always contain some degree of curvature. The dose impact of this may be significant.

  9. High Dose Rate Brachytherapy in Two 9 Gy Fractions in the Treatment of Locally Advanced Cervical Cancer - a South Indian Institutional Experience.

    Science.gov (United States)

    Ghosh, Saptarshi; Rao, Pamidimukkala Bramhananda; Kotne, Sivasankar

    2015-01-01

    Although 3D image based brachytherapy is currently the standard of treatment in cervical cancer, most of the centres in developing countries still practice orthogonal intracavitary brachytherapy due to financial constraints. The quest for optimum dose and fractionation schedule in high dose rate (HDR) intracavitary brachytherapy (ICBT) is still ongoing. While the American Brachytherapy Society recommends four to eight fractions of each less than 7.5 Gy, there are some studies demonstrating similar efficacy and comparable toxicity with higher doses per fraction. To assess the treatment efficacy and late complications of HDR ICBT with 9 Gy per fraction in two fractions. This is a prospective institutional study in Southern India carried on from 1st June 2012 to 31st July 2014. In this period, 76 patients of cervical cancer satisfying our inclusion criteria were treated with concurrent chemo-radiation following ICBT with 9 Gy per fraction in two fractions, five to seven days apart. The median follow-up period in the study was 24 months (range 10.6 - 31.2 months). The 2 year actuarial local control rate, disease-free survival and overall survival were 88.1%, 84.2% and 81.8% respectively. Although 38.2% patients suffered from late toxicity, only 3 patients had grade III late toxicity. In our experience, HDR brachytherapy with 9 Gy per fraction in two fractions is an effective dose fractionation for the treatment of cervical cancer with acceptable toxicity.

  10. American Society for Radiation Oncology (ASTRO) and American College of Radiology (ACR) Practice Guideline for the Performance of High-Dose-Rate Brachytherapy

    International Nuclear Information System (INIS)

    Erickson, Beth A.; Demanes, D. Jeffrey; Ibbott, Geoffrey S.; Hayes, John K.; Hsu, I-Chow J.; Morris, David E.; Rabinovitch, Rachel A.; Tward, Jonathan D.; Rosenthal, Seth A.

    2011-01-01

    High-Dose-Rate (HDR) brachytherapy is a safe and efficacious treatment option for patients with a variety of different malignancies. Careful adherence to established standards has been shown to improve the likelihood of procedural success and reduce the incidence of treatment-related morbidity. A collaborative effort of the American College of Radiology (ACR) and American Society for Therapeutic Radiation Oncology (ASTRO) has produced a practice guideline for HDR brachytherapy. The guideline defines the qualifications and responsibilities of all the involved personnel, including the radiation oncologist, physicist and dosimetrists. Review of the leading indications for HDR brachytherapy in the management of gynecologic, thoracic, gastrointestinal, breast, urologic, head and neck, and soft tissue tumors is presented. Logistics with respect to the brachytherapy implant procedures and attention to radiation safety procedures and documentation are presented. Adherence to these practice guidelines can be part of ensuring quality and safety in a successful HDR brachytherapy program.

  11. The irradiation tolerance dose of the proximal vagina

    International Nuclear Information System (INIS)

    Au, Samuel P.; Grigsby, Perry W.

    2003-01-01

    Purpose: The purpose of this investigation was to determine the irradiation tolerance level and complication rates of the proximal vagina to combined external irradiation and low dose rate (LDR) brachytherapy. Also, the mucosal tolerance for fractionated high dose rate (HDR) brachytherapy is further projected based on the biological equivalent dose (BED) of LDR for an acceptable complication rate. Materials and methods: Two hundred seventy-four patients with stages I-IV cervical carcinoma treated with irradiation therapy alone from 1987 to 1997 were retrospectively reviewed for radiation-associated late sequelae of the proximal vagina. All patients received LDR brachytherapy and 95% also received external pelvic irradiation. Follow-up ranged from 15 to 126 months (median, 43 months). The proximal vagina mucosa dose from a single ovoid (single source) or from both ovoids plus the tandem (all sources), together with the external irradiation dose, were used to derive the probability of a complication using the maximum likelihood logistic regression technique. The BED based on the linear-quadratic model was used to compute the corresponding tolerance levels for LDR or HDR brachytherapy. Results: Grades 1 and 2 complications occurred in 10.6% of patients and Grade 3 complications occurred in 3.6%. There were no Grade 4 complications. Complications occurred from 3 to 71 months (median, 7 months) after completion of irradiation, with over 60% occurring in the first year. By logistic regression analysis, both the mucosal dose from a single ovoid or that from all sources, combined with the external irradiation dose, demonstrate a statistically significant fit to the dose response complication curves (both with P=0.016). The single source dose was highly correlated with the all source dose with a cross-correlation coefficient 0.93. The all source dose was approximately 1.4 times the single source dose. Over the LDR brachytherapy dose rate range, the complication rate was

  12. SU-F-T-55: Reproducibility of Interstitial HDR Brachytherapy Plans

    Energy Technology Data Exchange (ETDEWEB)

    Lee, S; Ellis, R; Traughber, B; Podder, T [University Hospitals Case Medical Center, Cleveland, OH (United States)

    2016-06-15

    Purpose: Treating gynecological cancers with interstitial high-dose-rate (HDR) brachytherapy requires precise reconstruction of catheter positions to obtain accurate dosimetric plans. In this study, we investigated the degree of reproducibility of dosimetric plans for Syed HDR brachytherapy. Methods: We randomly selected five patients having cervix-vaginal cancer who were recently treated in our clinic with interstitial HDR brachytherapy with a prescription dose of 25–30 Gy in five fractions. Interstitial needles/catheters were placed under fluoroscopic guidance and intra-operative 3T MRI scan was performed to confirm the desired catheter placement for adequate target volume coverage. A CT scan was performed and fused with the MRI for delineating high-risk CTV (HR-CTV), intermediate-risk CTV (IR-CTV) and OARs. HDR treatment plans were generated using Oncentra planning software. A single plan was used for all five fractions of treatment for each patient. For this study, we took the original clinical plan and removed all the reconstructed catheters from the plan keeping the original contours unchanged. Then, we manually reconstructed all the catheters and entered the same dwell time from the first original clinical plan. The dosimetric parameters studied were: D90 for HR-CTV and IR-CV, and D2cc for bladder, rectum, sigmoid and bowel. Results: The mean of absolute differences in dosimetric coverage (D90) were (range): 1.3% (1.0–2.0%) and 2.0% (0.9–3.6%) for HR-CTV and IR-CTV, respectively. In case of OARs, the mean of absolute variations in D2cc were (range): 4.7% (0.7–8.9%) for bladder, 1.60% (0.3–3.2%) for rectum, 1.6% (0–3.9%) for sigmoid, and 1.8% (0–5.1%) for bowel. Conclusion: Overall, the reproducibility of interstitial HDR plans was within clinically acceptable limit. Observed maximum variation in D2cc for bladder. If number of catchers and dwell points were relatively low or any one catheter was heavily loaded, then reproducibility of the plan

  13. Late toxicity and five year outcomes after high-dose-rate brachytherapy as a monotherapy for localized prostate cancer

    International Nuclear Information System (INIS)

    Ghadjar, Pirus; Oesch, Sebastian L; Rentsch, Cyrill A; Isaak, Bernhard; Cihoric, Nikola; Manser, Peter; Thalmann, George N; Aebersold, Daniel M

    2014-01-01

    To determine the 5-year outcome after high-dose-rate brachytherapy (HDR-BT) as a monotherapy. Between 10/2003 and 06/2006, 36 patients with low (28) and intermediate (8) risk prostate cancer were treated by HDR-BT monotherapy. All patients received one implant and 4 fractions of 9.5 Gy within 48 hours for a total prescribed dose (PD) of 38 Gy. Five patients received concomitant androgen deprivation therapy (ADT). Toxicity was scored according to the common terminology criteria for adverse events from the National Cancer Institute (CTCAE) version 3.0. Biochemical recurrence was defined according to the Phoenix criteria and analyzed using the Kaplan Meier method. Predictors for late grade 3 GU toxicity were analyzed using univariate and multivariate Cox regression analyses. The median follow-up was 6.9 years (range, 1.5-8.0 years). Late grade 2 and 3 genitourinary (GU) toxicity was observed in 10 (28%) and 7 (19%) patients, respectively. The actuarial proportion of patients with late grade 3 GU toxicity at 5 years was 17.7%. Late grade 2 and 3 gastrointestinal (GI) toxicities were not observed. The crude erectile function preservation rate in patients without ADT was 75%. The 5 year biochemical recurrence-free survival (bRFS) rate was 97%. Late grade 3 GU toxicity was associated with the urethral volume (p = 0.001) and the urethral V 120 (urethral volume receiving ≥120% of the PD; p = 0.0005) after multivariate Cox regression. After HDR-BT monotherapy late grade 3 GU was observed relatively frequently and was associated with the urethral V 120 . GI toxicity was negligible, the erectile function preservation rate and the bRFS rate was excellent

  14. Accelerated partial-breast irradiation using high-dose-rate interstitial brachytherapy: 12-year update of a prospective clinical study

    International Nuclear Information System (INIS)

    Polgar, Csaba; Major, Tibor; Fodor, Janos; Sulyok, Zoltan; Somogyi, Andras; Loevey, Katalin; Nemeth, Gyoergy; Kasler, Miklos

    2010-01-01

    Background and purpose: To report the 12-year updated results of accelerated partial-breast irradiation (APBI) using multicatheter interstitial high-dose-rate (HDR) brachytherapy (BT). Patients and methods: Forty-five prospectively selected patients with T1N0-N1mi, nonlobular breast cancer without the presence of an extensive intraductal component and with negative surgical margins were treated with APBI after breast-conserving surgery (BCS) using interstitial HDR BT. A total dose of 30.3 Gy (n = 8) and 36.4 Gy (n = 37) in seven fractions within 4 days was delivered to the tumour bed plus a 1-2 cm margin. The median follow-up time was 133 months for surviving patients. Local and regional control, disease-free (DFS), cancer-specific (CSS), and overall survival (OS), as well as late side effects, and cosmetic results were assessed. Results: Four (8.9%) ipsilateral breast tumour recurrences were observed, for a 5-, 10-, and 12-year actuarial rate of 4.4%, 9.3%, and 9.3%, respectively. A total of two regional nodal failures were observed for a 12-year actuarial rate of 4.4%. The 12-year DFS, CSS, and OS was 75.3%, 91.1%, and 88.9%, respectively. Grade 3 fibrosis was observed in one patient (2.2%). No patient developed grade 3 teleangiectasia. Fat necrosis requiring surgical intervention occurred in one woman (2.2%). Cosmetic results were rated excellent or good in 35 patients (77.8%). Conclusions: Twelve-year results with APBI using HDR multicatheter interstitial implants continue to demonstrate excellent long-term local tumour control, survival, and cosmetic results with a low-rate of late side effects.

  15. Monte Carlo Dosimetry of the 60Co BEBIG High Dose Rate for Brachytherapy.

    Directory of Open Access Journals (Sweden)

    Luciana Tourinho Campos

    Full Text Available The use of high-dose-rate brachytherapy is currently a widespread practice worldwide. The most common isotope source is 192Ir, but 60Co is also becoming available for HDR. One of main advantages of 60Co compared to 192Ir is the economic and practical benefit because of its longer half-live, which is 5.27 years. Recently, Eckert & Ziegler BEBIG, Germany, introduced a new afterloading brachytherapy machine (MultiSource®; it has the option to use either the 60Co or 192Ir HDR source. The source for the Monte Carlo calculations is the new 60Co source (model Co0.A86, which is referred to as the new BEBIG 60Co HDR source and is a modified version of the 60Co source (model GK60M21, which is also from BEBIG.The purpose of this work is to obtain the dosimetry parameters in accordance with the AAPM TG-43U1 formalism with Monte Carlo calculations regarding the BEBIG 60Co high-dose-rate brachytherapy to investigate the required treatment-planning parameters. The geometric design and material details of the source was provided by the manufacturer and was used to define the Monte Carlo geometry. To validate the source geometry, a few dosimetry parameters had to be calculated according to the AAPM TG-43U1 formalism. The dosimetry studies included the calculation of the air kerma strength Sk, collision kerma in water along the transverse axis with an unbounded phantom, dose rate constant and radial dose function. The Monte Carlo code system that was used was EGSnrc with a new cavity code, which is a part of EGS++ that allows calculating the radial dose function around the source. The spectrum to simulate 60Co was composed of two photon energies, 1.17 and 1.33 MeV. Only the gamma part of the spectrum was used; the contribution of the electrons to the dose is negligible because of the full absorption by the stainless-steel wall around the metallic 60Co. The XCOM photon cross-section library was used in subsequent simulations, and the photoelectric effect, pair

  16. Under-utilisation of high-dose-rate brachytherapy boost in men with intermediate-high risk prostate cancer treated with external beam radiotherapy.

    Science.gov (United States)

    Ong, Wee Loon; Evans, Sue M; Millar, Jeremy L

    2018-04-01

    The aim of this study was to evaluate the use of high-dose-rate brachytherapy (HDR-BT) boost with definitive external beam radiotherapy (EBRT) in prostate cancer (CaP) management. The study population comprised men with intermediate-high risk CaP captured in the population-based Prostate Cancer Outcome Registry Victoria (PCOR-Vic), treated with EBRT from January 2010 to December 2015. The primary outcome is the proportion of men who received HDR-BT boost. Multivariate logistic regressions were used to evaluate the effect of patient-, tumour- and treatment-factors on the likelihood of HDR-BT use. Medicare Benefit Schedule (MBS) data was accessed to evaluate the Australia-wide pattern of HDR-BT use. One thousand eight hundred and six patients were included in this study - 886 (49%) intermediate-risk, and 920 (51%) high-risk CaP patients. Overall, only 124 (7%) patients had EBRT + HDR-BT - 47 (5%) intermediate-risk and 77 (8%) high-risk CaP patients (P = 0.01). There is higher proportion of patients who had HDR-BT in public institutions (7% public vs. 3% private, P = 0.005) and in metropolitan centres (9% metropolitan vs. 2% regional, P Victorian men with CaP. The decline in HDR-BT use was also observed nationally. © 2017 The Royal Australian and New Zealand College of Radiologists.

  17. High dose-rate brachytherapy source position quality assurance using radiochromic film

    International Nuclear Information System (INIS)

    Evans, M.D.C.; Devic, S.; Podgorsak, E.B.

    2007-01-01

    Traditionally, radiographic film has been used to verify high-dose-rate brachytherapy source position accuracy by co-registering autoradiographic and diagnostic images of the associated applicator. Filmless PACS-based clinics that do not have access to radiographic film and wet developers may have trouble performing this quality assurance test in a simple and practical manner. We describe an alternative method for quality assurance using radiochromic-type film. In addition to being easy and practical to use, radiochromic film has some advantages in comparison with traditional radiographic film when used for HDR brachytherapy quality assurance

  18. A multicentre ‘end to end’ dosimetry audit for cervix HDR brachytherapy treatment

    International Nuclear Information System (INIS)

    Palmer, Antony L.; Diez, Patricia; Gandon, Laura; Wynn-Jones, Andrea; Bownes, Peter; Lee, Chris; Aird, Edwin; Bidmead, Margaret; Lowe, Gerry; Bradley, David; Nisbet, Andrew

    2015-01-01

    Purpose: To undertake the first multicentre fully ‘end to end’ dosimetry audit for HDR cervix brachytherapy, comparing planned and delivered dose distributions around clinical treatment applicators, with review of local procedures. Materials and methods: A film-dosimetry audit was performed at 46 centres, including imaging, applicator reconstruction, treatment planning and delivery. Film dose maps were calculated using triple-channel dosimetry and compared to RTDose data from treatment planning systems. Deviations between plan and measurement were quantified at prescription Point A and using gamma analysis. Local procedures were also discussed. Results: The mean difference between planned and measured dose at Point A was −0.6% for plastic applicators and −3.0% for metal applicators, at standard uncertainty 3.0% (k = 1). Isodose distributions agreed within 1 mm over a dose range 2–16 Gy. Mean gamma passing rates exceeded 97% for plastic and metal applicators at 3% (local) 2 mm criteria. Two errors were found: one dose normalisation error and one applicator library misaligned with the imaged applicator. Suggestions for quality improvement were also made. Conclusions: The concept of ‘end to end’ dosimetry audit for HDR brachytherapy has been successfully implemented in a multicentre environment, providing evidence that a high level of accuracy in brachytherapy dosimetry can be achieved

  19. Radical prostatectomy vs high-dose-rate brachytherapy for prostate cancer. Health-related quality-of-life effects

    International Nuclear Information System (INIS)

    Jo, Yoshimasa; Fujisawa, Masato

    2004-01-01

    A screening comparison was made of health-related quality-of-life (HRQOL) outcomes between two primary treatment modalities for localized prostate cancer: radical prostatectomy (RP) and iridium-192 (Ir-192) high-dose rate brachytherapy (HDR-BT) followed by external beam radiotherapy (EBRT). The subjects were 182 patients diagnosed with T1c to T3bN0M0 prostate cancer between October 1997 through August 2002 who underwent RP (n=89) or HDR-BT with 36.8 Gy of EBRT (n=93) and follow-up for at least 6 months. A postal survey was sent, in which HRQOL was assessed using the Short Form Health Survey (SF-36), and disease-specific QOL using the University of California Los Angeles Prostate Cancer Index (UCLA-PCI). We obtained responses to questionnaires from 151 out of 182 patients (83.0%; RP, 78.7%, HDR-BT, 87.1%). No significant difference was observed in SF-36 scale scores between RP and HDR-BT. In the UCLA-PCI, the HDR-BT group had better urinary function (UF, p<0.001) and sexual function (SF, p=0.0272), whereas the RP group had better bowel bother (BB, p=0.0425). In patients with at least 2 years of follow-up, UF (p<0.001) and sexual bother (SB, p=0.0286) were better for the HDR-BT group than for the RP group. HDR-BT patients had significantly better UF (p=0.009) and SB (p=0.0134) than even patients with uni-lateral nerve-sparing RP (n=30). When planning treatment, QOL concerns including mental health issues associated with prostate cancer need to be addressed with the patients, as well as the potential side effects. (author)

  20. Novel use of the Contura for high dose rate cranial brachytherapy.

    Science.gov (United States)

    Scanderbeg, Daniel J; Alksne, John F; Lawson, Joshua D; Murphy, Kevin T

    2011-01-01

    A popular choice for treatment of recurrent gliomas was cranial brachytherapy using the GliaSite Radiation Therapy System. However, this device was taken off the market in late 2008, thus leaving a treatment void. This case study presents our experience treating a cranial lesion for the first time using a Contura multilumen, high-dose-rate (HDR) brachytherapy balloon applicator. The patient was a 47-year-old male who was diagnosed with a recurrent right frontal anaplastic oligodendroglioma. Previous radiosurgery made him a good candidate for brachytherapy. An intracavitary HDR balloon brachytherapy device (Contura) was placed in the resection cavity and treated with a single fraction of 20 Gy. The implant, treatment, and removal of the device were all completed without incident. Dosimetry of the device was excellent because the dose conformed very well to the target. V90, V100, V150, and V200 were 98.9%, 95.7%, 27.2, and 8.8 cc, respectively. This patient was treated successfully using the Contura multilumen balloon. Contura was originally designed for deployment in a postlumpectomy breast for treatment by accelerated partial breast irradiation. Being an intracavitary balloon device, its similarity to the GliaSite system makes it a viable replacement candidate. Multiple lumens in the device also make it possible to shape the dose delivered to the target, something not possible before with the GliaSite applicator. Copyright © 2011 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

  1. Preliminary results of study comparing HDR with LDR brachytherapy for IIIb cervical cancer

    International Nuclear Information System (INIS)

    Trippe, N.; Pellizzon, A.C.A.; Novaes, P.; Salvajoli, J.V.; Fogaroli, R.; Maia, M.A.C.; Baraldi, H.; Ferrigno, R.

    1996-01-01

    Since 1992 we have been using a Micro-Selectron HDR device, working with Iridium 192 to treat the cervical cancer and some others pathologies. With a minimum follow up of 24 months, 59 patients with cervical cancer were randomizated for one of the following schedule of treatment: EBRT - 45Gy - fx 1,8Gy plus Brachytherapy 1-HDR - 36 (61%) - 4 weekly insertions of 6,0Gy at point A 2-LDR - 29 (39%) - two insertions fifteen days apart of 17,5Gy at point A EBRT was performed with a Linac 4MV, in box arrangement and parametrial complementation of dose with AP-PA fields. For Brachytherapy Fletcher Colpostats are used in association with intrauterine tamdens, in both arms. Brachyterapy starts in HDR group after ten days of the beginning of the treatment. The total time of treatment is shortened here in two weeks. LDR brachytherapy starts only after the end of EBRT. Results - local control was 61% in 12 months and 50% in 24 months for HDR group, versus 52,6% and 47,8% for LDR group. Local failures of 39% and 50% in 12 and 24 months for HDR and 47,8% and 52,8% for LDR groups respectively. Complications were restricted to rectites and cistites - 8,3% for HDR and 13% for LDR. Conclusions - HDR brachytherapy has an equivalent local control when compared to LDR, can treat a larger number of patients in a shorter period, has possibilities of dose optimizations and decrease the radiation exposure to the staff

  2. High-dose rate intra-operative radiation therapy for local advanced and recurrent colorectal cancer

    International Nuclear Information System (INIS)

    Harrison, L.B.; Mychalczak, B.; Enker, W.; Anderson, L.; Cohen, A.E.; Minsky, B.

    1996-01-01

    In an effort to improve the local control for advanced and recurrent cancers of the rectum, we have integrated high-dose rate intra-operative radiation therapy (HDR-IORT) into the treatment program. Between 11/92 and 10/95, 47 patients (pts) were treated. There were 26 males and 21 females whose ages ranged from 30-80 (median = 62) years. There were 19 pts with primary unresectable rectal cancer, and 28 pts who were treated for recurrent rectal cancer. Histology was adenocarcinoma - 45 pts, squamous cancer - 2 pts. The range of follow-up is 1-34 months (median = 14 months). The majority of primary unresectable pts received pre-operative radiation therapy (4500-5040 cGy) with chemotherapy (5-FU with Leucovorin) 4-6 weeks later, they underwent resection + HDR-IORT (1200 cGy). For the 28 pts with recurrent cancer, the majority received surgery and HDR-IORT alone because they had received prior RT. For the pts with primary unresectable disease, actuarial 2-year local control was 77%, actuarial distant metastasis-free survival was 71%, disease free survival was 66%, and overall survival was 84%. For those pts with recurrent disease, actuarial 2-year local control rate was 65%, distant metastasis-free survival was 65%, disease free survival was 47%, and overall survival was 61%. Complications occurred in 36%. There were no cases where the anatomical distribution of disease, or technical limitations prevented the adequate delivery of HDR-IORT. We conclude that this technique was most versatile, and enabled all appropriate pts to receive IORT. The preliminary data in terms of local control are encouraging, even for the poor prognostic sub-group of pts with recurrent cancer

  3. Time Course and Accumulated Risk of Severe Urinary Adverse Events After High- Versus Low-Dose-Rate Prostate Brachytherapy With or Without External Beam Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Tward, Jonathan D., E-mail: Jonathan.Tward@hci.utah.edu [Huntsman Cancer Institute, University of Utah, Salt Lake City, Utah (United States); Jarosek, Stephanie; Chu, Haitao [University of Minnesota, Minneapolis, Minnesota (United States); Thorpe, Cameron; Shrieve, Dennis C. [Huntsman Cancer Institute, University of Utah, Salt Lake City, Utah (United States); Elliott, Sean [University of Minnesota, Minneapolis, Minnesota (United States)

    2016-08-01

    Purpose: Severe urinary adverse events (UAEs) include surgical treatment of urethral stricture, urinary incontinence, and radiation cystitis. We compared the incidence of grade 3 UAEs, according to the Common Terminology Criteria for Adverse Events, after low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy, as well as after LDR plus external beam radiation therapy (EBRT) and HDR plus EBRT. Methods and Materials: Men aged >65 years with nonmetastatic prostate cancer were identified from the Surveillance, Epidemiology, and End Results–Medicare database who were treated with LDR (n=12,801), HDR (n=685), LDR plus EBRT (n=8518), or HDR plus EBRT (n=2392). The populations were balanced by propensity weighting, and the Kaplan-Meier incidence of severe UAEs was compared. Propensity-weighted Cox proportional hazards models were used to compare the adjusted hazard of UAEs. These UAEs were compared with those in a cohort of men not treated for prostate cancer. Results: Median follow-up was 4.3 years. At 8 years, the propensity-weighted cumulative UAE incidence was highest after HDR plus EBRT (26.6% [95% confidence interval, 23.8%-29.7%]) and lowest after LDR (15.7% [95% confidence interval, 14.8%-16.6%]). The absolute excess risk over nontreated controls at 8 years was 1.9%, 3.8%, 8.4%, and 12.9% for LDR, HDR, LDR plus EBRT, and HDR plus EBRT, respectively. These represent numbers needed to harm of 53, 26, 12, and 8 persons, respectively. The additional risk of development of a UAE related to treatment for LDR, LDR plus EBRT, and HDR plus EBRT was greatest within the 2 years after treatment and then continued to decline over time. Beyond 4 years, the risk of development of a new severe UAE matched the baseline risk of the control population for all treatments. Conclusions: Toxicity differences were observed between LDR and HDR, but the differences did not meet statistical significance. However, combination radiation therapy (either HDR plus EBRT or LDR plus

  4. Time Course and Accumulated Risk of Severe Urinary Adverse Events After High- Versus Low-Dose-Rate Prostate Brachytherapy With or Without External Beam Radiation Therapy

    International Nuclear Information System (INIS)

    Tward, Jonathan D.; Jarosek, Stephanie; Chu, Haitao; Thorpe, Cameron; Shrieve, Dennis C.; Elliott, Sean

    2016-01-01

    Purpose: Severe urinary adverse events (UAEs) include surgical treatment of urethral stricture, urinary incontinence, and radiation cystitis. We compared the incidence of grade 3 UAEs, according to the Common Terminology Criteria for Adverse Events, after low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy, as well as after LDR plus external beam radiation therapy (EBRT) and HDR plus EBRT. Methods and Materials: Men aged >65 years with nonmetastatic prostate cancer were identified from the Surveillance, Epidemiology, and End Results–Medicare database who were treated with LDR (n=12,801), HDR (n=685), LDR plus EBRT (n=8518), or HDR plus EBRT (n=2392). The populations were balanced by propensity weighting, and the Kaplan-Meier incidence of severe UAEs was compared. Propensity-weighted Cox proportional hazards models were used to compare the adjusted hazard of UAEs. These UAEs were compared with those in a cohort of men not treated for prostate cancer. Results: Median follow-up was 4.3 years. At 8 years, the propensity-weighted cumulative UAE incidence was highest after HDR plus EBRT (26.6% [95% confidence interval, 23.8%-29.7%]) and lowest after LDR (15.7% [95% confidence interval, 14.8%-16.6%]). The absolute excess risk over nontreated controls at 8 years was 1.9%, 3.8%, 8.4%, and 12.9% for LDR, HDR, LDR plus EBRT, and HDR plus EBRT, respectively. These represent numbers needed to harm of 53, 26, 12, and 8 persons, respectively. The additional risk of development of a UAE related to treatment for LDR, LDR plus EBRT, and HDR plus EBRT was greatest within the 2 years after treatment and then continued to decline over time. Beyond 4 years, the risk of development of a new severe UAE matched the baseline risk of the control population for all treatments. Conclusions: Toxicity differences were observed between LDR and HDR, but the differences did not meet statistical significance. However, combination radiation therapy (either HDR plus EBRT or LDR plus

  5. Time Course and Accumulated Risk of Severe Urinary Adverse Events After High- Versus Low-Dose-Rate Prostate Brachytherapy With or Without External Beam Radiation Therapy.

    Science.gov (United States)

    Tward, Jonathan D; Jarosek, Stephanie; Chu, Haitao; Thorpe, Cameron; Shrieve, Dennis C; Elliott, Sean

    2016-08-01

    Severe urinary adverse events (UAEs) include surgical treatment of urethral stricture, urinary incontinence, and radiation cystitis. We compared the incidence of grade 3 UAEs, according to the Common Terminology Criteria for Adverse Events, after low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy, as well as after LDR plus external beam radiation therapy (EBRT) and HDR plus EBRT. Men aged >65 years with nonmetastatic prostate cancer were identified from the Surveillance, Epidemiology, and End Results-Medicare database who were treated with LDR (n=12,801), HDR (n=685), LDR plus EBRT (n=8518), or HDR plus EBRT (n=2392). The populations were balanced by propensity weighting, and the Kaplan-Meier incidence of severe UAEs was compared. Propensity-weighted Cox proportional hazards models were used to compare the adjusted hazard of UAEs. These UAEs were compared with those in a cohort of men not treated for prostate cancer. Median follow-up was 4.3 years. At 8 years, the propensity-weighted cumulative UAE incidence was highest after HDR plus EBRT (26.6% [95% confidence interval, 23.8%-29.7%]) and lowest after LDR (15.7% [95% confidence interval, 14.8%-16.6%]). The absolute excess risk over nontreated controls at 8 years was 1.9%, 3.8%, 8.4%, and 12.9% for LDR, HDR, LDR plus EBRT, and HDR plus EBRT, respectively. These represent numbers needed to harm of 53, 26, 12, and 8 persons, respectively. The additional risk of development of a UAE related to treatment for LDR, LDR plus EBRT, and HDR plus EBRT was greatest within the 2 years after treatment and then continued to decline over time. Beyond 4 years, the risk of development of a new severe UAE matched the baseline risk of the control population for all treatments. Toxicity differences were observed between LDR and HDR, but the differences did not meet statistical significance. However, combination radiation therapy (either HDR plus EBRT or LDR plus EBRT) increases the risk of severe UAEs compared with HDR

  6. Physics and quality assurance for brachytherapy - Part I: High dose rates

    International Nuclear Information System (INIS)

    Anderson, Lowell L.

    1997-01-01

    Purpose: To review the physical aspects of high dose rate (HDR) brachytherapy, including commissioning and quality assurance, source calibration and dose distribution measurements, and treatment planning methods. Following the introduction of afterloading in brachytherapy, development efforts to make it 'remote' culminated in 1964 with the near-simultaneous appearance of remote afterloaders in five major medical centers. Four of these machines were 'high dose rate', three employing 60Co and one (the GammaMed) using a single, cable-mounted 192Ir source. Stepping-motor source control was added to the GammaMed in 1974, making it the precursor of modern remote afterloaders, which are now suitable for interstitial, well as intracavitary brachytherapy by virtue of small source-diameter and indexer-accessed multiple channels. Because the 192Ir sources currently used in HDR remote afterloaders are supplied at a nominal air-kerma strength of 11.4 cGy cm2 s-1 (10 Ci), are not collimated in clinical use, and emit a significant fraction (15%) of photons at energies greater than 600 keV, shielding and facility design must be undertaken as carefully and thoroughly as for external beam installations. Licensing requirements of regulatory agencies must be met with respect both to maximum permissible dose limits and to the existence and functionality of safety devices (door interlocks, radiation monitors, etc.). Commissioning and quality assurance procedures that must be documented for HDR remote afterloading relate to (1) machine, applicator, guide-tube, and facility functionality checks, (2) source calibration, (3) emergency response readiness, (4) planning software evaluation, and (5) independent checks of clinical dose calculations. Source calibration checks must be performed locally, either by in-air measurement of air kerma strength or with a well ionization chamber calibrated (by an accredited standards laboratory) against an in-air measurement of air kerma strength for the

  7. Adjuvant high dose rate vaginal cuff brachytherapy for early stage endometrial cancer

    International Nuclear Information System (INIS)

    Tannehill, S.P.; Petereit, D.G.; Schink, J.C.; Grosen, E.A.; Hartenbach, E.M.; Thomadsen, B.R.; Buchler, D.A.

    1997-01-01

    Objective: To determine the efficacy and complications of adjuvant high dose rate (HDR) vaginal cuff brachytherapy (VCB) in patients (pts) with low risk endometrial carcinoma. Materials and Methods: Since 1989, 154 patients were treated with outpatient adjuvant VCB for low risk endometrial cancer (Stage IA-14%, Stage IB-82%). Four percent of patients with stage IC disease were treated with VCB only because of medical contraindications to pelvic radiation. Patients had the following histologic grades: 53% grade 1, 40% grade 2, 5% grade 3 and 3% unknown (99%-adenocarcinoma, <1% papillary serous histology). Seventy-three percent of patients had their surgery (TAH-BSO) performed at an outside institution with minimal surgical assessment of the lymph nodes. At a median of 6 weeks after surgery, patients were treated with 2 HDR VCB insertions delivered 1 week apart. Ovoids were placed at the vaginal apex to deliver 16.2 Gy per fraction to the vaginal surface (LDR equivalent of 60 Gy at 100 cGy/h) under conscious outpatient sedation. All clinical endpoints were calculated using the Kaplan Meier method. Complications were scored using the RTOG 5-tiered system. Results: The median time in the brachytherapy suite was 60 minutes with no acute complications observed. With a median follow-up of 33 months (3-79 months), the 4-year overall and disease-free survival were 93% and 96% respectively. Five patients (3%) recurred: 2 intra-abdominally, 1 with lung metastases, and 2 in the pelvic lymph nodes. There were no vaginal cuff recurrences. The single patient with an isolated pelvic sidewall recurrence was salvaged with pelvic RT. Six patients developed a small area of asymptomatic necrosis at the vaginal cuff, which spontaneously healed at a median time of 4 months. There were no grade 3 or greater late tissue toxicities. No patient experienced significant vaginal stenosis, with 20% of the patients experiencing mild fibrosis of the vaginal apex. Conclusions: Adjuvant HDR VCB in 2

  8. Local vaginal anesthesia during high-dose-rate intracavitary brachytherapy for cervical cancer

    International Nuclear Information System (INIS)

    Chen, H.-C.; Wan Leung, Stephen; Wang, C.-J.; Sun, L.-M.; Fang, F.-M.; Huang, E.-Y.; Wang, S.-J.; Yang, C.-W.

    1998-01-01

    Purpose: To evaluate the clinical efficacy of local vaginal lidocaine application for pain relief during high-dose-rate (HDR) intracavitary brachytherapy for patients with cervical cancer, and to investigate sequential changes in serum levels of lidocaine during the procedures. Methods and Materials: This prospective study was designed to examine the analgesic effect, physical response, and side effects of local anesthesia during HDR intracavitary brachytherapy. Forty patients were enrolled. All patients received 10-15 MV X-rays to the pelvis with a total dose of 45-59.4 Gy 5-6 weeks before undergoing HDR intracavitary brachytherapy. All patients underwent first intracavitary brachytherapy under general anesthesia. These patients were randomly allocated to receive one of two different treatment protocols as follows: (1) treatment session - control session - treatment session - control session; or (2) control session - treatment session- control session - treatment session. In the treatment sessions, topical anesthesia was administered using 4 ml of 10% lidocaine solution sprayed liberally on the cervix and vagina during intracavitary brachytherapy. In the control sessions, a placebo was administered in the same manner during brachytherapy. The Hensche's applicators for brachytherapy were inserted into the cervix and vagina 5 min after lidocaine application. The visual analogue scale (VAS) was used to assess pain and discomfort during brachytherapy. Blood pressure and heart rates were measured to evaluate the physiological response. Another prospective study was then performed to investigate the sequential changes of serum lidocaine levels during the anesthetic procedure. Eleven additional patients with similar disease state and demographic characteristics were enrolled and blood samples were obtained before, and 5, 15, 30, and 45 min after the initiation of lidocaine application. Results: The mean VAS values recorded during the treatment sessions and control

  9. The use of high-dose-rate brachytherapy alone after lumpectomy in patients with early-stage breast cancer treated with breast-conserving therapy

    International Nuclear Information System (INIS)

    Baglan, Kathy L.; Martinez, Alvaro A.; Frazier, Robert C.; Kini, Vijay R.; Kestin, Larry L.; Chen, Peter Y.; Edmundson, Greg; Mele, Elizabeth; Jaffray, David; Vicini, Frank A.

    2001-01-01

    Purpose: We present the preliminary results of our in-house protocol using outpatient high-dose-rate (HDR) brachytherapy as the sole radiation modality following lumpectomy in patients with early-stage breast cancer. Methods and Materials: Thirty-seven patients with 38 Stage I-II breast cancers received radiation to the lumpectomy cavity alone using an HDR interstitial implant with 192 Ir. A minimum dose of 32 Gy was delivered on an outpatient basis in 8 fractions of 4 Gy to the lumpectomy cavity plus a 1- to 2-cm margin over consecutive 4 days. Results: Median follow-up is 31 months. There has been one ipsilateral breast recurrence for a crude failure rate of 2.6% and no regional or distant failures. Wound healing was not impaired in patients undergoing an open-cavity implant. Three minor breast infections occurred, and all resolved with oral antibiotics. The cosmetic outcome was good to excellent in all patients. Conclusion: In selected patients with early-stage breast cancer, treatment of the lumpectomy cavity alone with outpatient HDR brachytherapy is both technically feasible and well tolerated. Early results are encouraging, however, longer follow-up is necessary before equivalence to standard whole-breast irradiation can be established and to determine the most optimal radiation therapy technique to be employed

  10. Poster - Thur Eve - 03: LDR to HDR: RADPOS applications in brachytherapy.

    Science.gov (United States)

    Cherpak, A J; Cygler, J E; Kertzscher, G; E, C; Perry, G

    2012-07-01

    The RADPOS in vivo dosimetry system combines an electromagnetic positioning sensor and either one or five MOSFET dosimeters. The feasibility of using the system for quality control has been explored for a range of radiotherapy treatment techniques including most recently transperineal interstitial permanent prostate brachytherapy and high dose rate (HDR) treatments. Dose and position information was collected by a RADPOS array detector inside a Foley catheter within patients' urethra during permanent seed implantation. Ten patients were studied, and average displacement during implantation was Δr = (1.4-5.1) mm, with movements up to 9.7 mm due to the removal of the transrectal ultrasound probe. Maximum integral dose in the prostatic urethra ranged from 110-195 Gy, and it was found that the dose can change up to 63 cGy (62.0%) depending on whether the rectal probe is in place. For HDR, a RADPOS detector was first calibrated with an Ir-192 source. A treatment was then simulated using a total of 50 dwell positions in 5 catheters in an acrylic phantom. Dwell positions ranged from 1 to 10 cm away from the RADPOS detector and dose was measured for each source position. An average calibration coefficient of 0.74±0.11 cGy/mV was calculated for the detector and the average absolute difference between measured values and expected dose was 0.7±5.4 cGy (5±20%). The demonstrated accuracy of RADPOS dose measurements along with its ability to simultaneously measure displacement makes it a powerful tool for brachytherapy treatments, where high dose gradients can present unique in vivo dosimetry challenges. © 2012 American Association of Physicists in Medicine.

  11. Inverse planning anatomy-based dose optimization for HDR-brachytherapy of the prostate using fast simulated annealing algorithm and dedicated objective function

    International Nuclear Information System (INIS)

    Lessard, Etienne; Pouliot, Jean

    2001-01-01

    An anatomy-based dose optimization algorithm is developed to automatically and rapidly produce a highly conformal dose coverage of the target volume while minimizing urethra, bladder, and rectal doses in the delivery of an high dose-rate (HDR) brachytherapy boost for the treatment of prostate cancer. The dwell times are optimized using an inverse planning simulated annealing algorithm (IPSA) governed entirely from the anatomy extracted from a CT and by a dedicated objective function (cost function) reflecting clinical prescription and constraints. With this inverse planning approach, the focus is on the physician's prescription and constraint instead of on the technical limitations. Consequently, the physician's control on the treatment is improved. The capacity of this algorithm to represent the physician's prescription is presented for a clinical prostate case. The computation time (CPU) for IPSA optimization is less than 1 min (41 s for 142 915 iterations) for a typical clinical case, allowing fast and practical dose optimization. The achievement of highly conformal dose coverage to the target volume opens the possibility to deliver a higher dose to the prostate without inducing overdosage of urethra and normal tissues surrounding the prostate. Moreover, using the same concept, it will be possible to deliver a boost dose to a delimited tumor volume within the prostate. Finally, this method can be easily extended to other anatomical sites

  12. A quantitative analysis of two-dimensional manually segmented transrectal ultrasound axial images in planning high dose rate brachytherapy for prostate cancer

    Directory of Open Access Journals (Sweden)

    Dabić-Stanković Kata

    2017-01-01

    Full Text Available Background/Aim. Prostate delineation, pre-planning and catheter implantation procedures, in high-dose rate brachytherapy (HDR-BT, are commonly based on the prostate manually segmented transrectal ultrasound (TRUS images. The aim of this study was to quantitatively analyze the consistency of prostate capsule delineation, done by a single therapist, prior to each HDR-BT fraction and the changes in the shape of the prostate capsule during HDR-BT, using two dimensional (2D TRUS axial image. Methods. A group of 16 patients were treated at the Medical System Belgrade Brachytherapy Department with definitive HDRBT. The total applied median dose of 52 Gy was divided into four individual fractions, each fraction being delivered 2– 3 weeks apart. Real time prostate axial visualization and the manual segmentation prior to each fraction were performed using B-K Medical ultrasound. Quantitative analyses, analysis of an area and shape were applied on 2D-TRUS axial images of the prostate. Area analyses were used to calculate the average value of the cross-sectional area of the prostate image. The parameters of the prostate shape, the fractal dimension and the circularity ratio of the prostate capsule contour were estimated at the maximum axial cross section of the prostate image. Results. The sample group consisted of four phases, each phase being performed prior to the first, second, third and fourth HDR-BT fraction, respectively. Statistical analysis showed that during HDR-BT fractions there were no significant differences in the average value of area, as well as in the maximum shape of prostate capsule. Conclusions. Quantitative analysis of TRUS axial prostate segmented images shows a successful capsule delineation in the series of manually segmented TRUS images, and the prostate maximum shape remaining unchanged during HDR-BT fractions.

  13. The Meaning and Experience of Patients Undergoing Rectal High-Dose-Rate Brachytherapy.

    Science.gov (United States)

    Perez, Samara; Néron, Sylvain; Benc, Renata; Rosberger, Zeev; Vuong, Té

    2016-01-01

    High-dose-rate (HDR) brachytherapy is a precise form of radiation therapy that targets cancerous tumors by directly applying the radiation source at the site or directly next to the tumor. Patients often experience but underreport pain and anxiety related to cancer treatments. At present, there is no research available concerning the pervasiveness and intensity of patients' pain and anxiety during rectal brachytherapy. The aim of this study was to examine patients' thoughts, emotions, coping strategies, physical sensations, and needs during rectal HDR brachytherapy treatment. Twenty-five patients with rectal cancer were interviewed using a semi-structured qualitative interview following the completion of their brachytherapy treatment delivered at a Montreal-based hospital in Quebec, Canada. The experiences of pain and discomfort varied greatly between patients and were linked to the meaning patients attributed to the treatment itself, sense of time, the body's lithotomic position, insertion of the treatment applicator, and the patients' sense of agency and empowerment during the procedure. Patients drew upon a variety of internal and external resources to help them cope with discomfort. Staff need to know about the variation in the physical and emotional experiences of patients undergoing this treatment. Clinical teams can tailor their procedural behavior (eg, using certain language, psychosocial interventions) according to patients' needs to increase patients' comfort and ultimately improve their experience of HDR rectal brachytherapy.

  14. Comparison of leak opening and leak rate calculations to HDR experimental results

    International Nuclear Information System (INIS)

    Grebner, H.; Hoefler, A.; Hunger, H.

    1993-01-01

    During the last years a number of calculations of leak opening and leak rate for through cracks in piping components have been performed. Analyses are pre- or mostly post-calculations to experiments performed at the HDR facility under PWR operating conditions. Piping components under consideration were small diameter straight pipes with circumferential cracks, pipe bends with longitudinal or circumferential cracks and pipe branches with weldment cracks. The components were loaded by internal pressure and opening as well as closing bending moment. The finite element method and two-phase flow leak rate programs were used for the calculations. Results of the analyses are presented as J-integral values, crack opening displacements and areas and leak rates as well as comparisons to the experimental results

  15. Interfraction patient motion and implant displacement in prostate high dose rate brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Fox, C. D.; Kron, T.; Leahy, M.; Duchesne, G.; Williams, S.; Tai, K. H.; Haworth, A.; Herschtal, A.; Foroudi, F. [Department of Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Nursing Service, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Department of Radiation Oncology, Peter MacCallum Cancer Centre and University of Melbourne, Melbourne, Victoria 3002 (Australia); Department of Physical Sciences, Peter MacCallum Cancer Centre and Royal Melbourne Insititute of Technology, Melbourne, Victoria 3000 (Australia); Department of Biostatistics and Clinical Trials, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Department of Radiation Oncology, Peter MacCallum Cancer Centre and University of Melbourne, Melbourne, Victoria 3010 (Australia)

    2011-11-15

    Purpose: To quantify movement of prostate cancer patients undergoing treatment, using an in-house developed motion sensor in order to determine a relationship between patient movement and high dose rate (HDR) brachytherapy implant displacement. Methods: An electronic motion sensor was developed based on a three axis accelerometer. HDR brachytherapy treatment for prostate is delivered at this institution in two fractions 24 h apart and 22 patients were monitored for movement over the interval between fractions. The motion sensors functioned as inclinometers, monitoring inclination of both thighs, and the inclination and roll of the abdomen. The implanted HDR brachytherapy catheter set was assessed for displacement relative to fiducial markers in the prostate. Angle measurements and angle differences over a 2 s time base were binned, and the standard deviations of the resulting frequency distributions used as a metric for patient motion in each monitored axis. These parameters were correlated to measured catheter displacement using regression modeling. Results: The mean implant displacement was 12.6 mm in the caudal direction. A mean of 19.95 h data was recorded for the patient cohort. Patients generally moved through a limited range of angles with a mean of the exception of two patients who spent in excess of 2 h lying on their side. When tested for a relationship between movement in any of the four monitored axes and the implant displacement, none was significant. Conclusions: It is not likely that patient movement influences HDR prostate implant displacement. There may be benefits to patient comfort if nursing protocols were relaxed to allow patients greater freedom to move while the implant is in situ.

  16. Interfraction patient motion and implant displacement in prostate high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Fox, C. D.; Kron, T.; Leahy, M.; Duchesne, G.; Williams, S.; Tai, K. H.; Haworth, A.; Herschtal, A.; Foroudi, F.

    2011-01-01

    Purpose: To quantify movement of prostate cancer patients undergoing treatment, using an in-house developed motion sensor in order to determine a relationship between patient movement and high dose rate (HDR) brachytherapy implant displacement. Methods: An electronic motion sensor was developed based on a three axis accelerometer. HDR brachytherapy treatment for prostate is delivered at this institution in two fractions 24 h apart and 22 patients were monitored for movement over the interval between fractions. The motion sensors functioned as inclinometers, monitoring inclination of both thighs, and the inclination and roll of the abdomen. The implanted HDR brachytherapy catheter set was assessed for displacement relative to fiducial markers in the prostate. Angle measurements and angle differences over a 2 s time base were binned, and the standard deviations of the resulting frequency distributions used as a metric for patient motion in each monitored axis. These parameters were correlated to measured catheter displacement using regression modeling. Results: The mean implant displacement was 12.6 mm in the caudal direction. A mean of 19.95 h data was recorded for the patient cohort. Patients generally moved through a limited range of angles with a mean of the exception of two patients who spent in excess of 2 h lying on their side. When tested for a relationship between movement in any of the four monitored axes and the implant displacement, none was significant. Conclusions: It is not likely that patient movement influences HDR prostate implant displacement. There may be benefits to patient comfort if nursing protocols were relaxed to allow patients greater freedom to move while the implant is in situ.

  17. Interstitial high-dose-rate brachytherapy boost: The feasibility and cosmetic outcome of a fractionated outpatient delivery scheme

    International Nuclear Information System (INIS)

    Manning, Matthew A.; Arthur, Douglas W.; Schmidt-Ullrich, Rupert K.; Arnfield, Mark R.; Amir, Cyrus; Zwicker, Robert D.

    2000-01-01

    Purpose: To evaluate the feasibility, potential toxicity, and cosmetic outcome of fractionated interstitial high dose rate (HDR) brachytherapy boost for the management of patients with breast cancer at increased risk for local recurrence. Methods and Materials: From 1994 to 1996, 18 women with early stage breast cancer underwent conventionally fractionated whole breast radiotherapy (50-50.4 Gy) followed by interstitial HDR brachytherapy boost. All were considered to be at high risk for local failure. Seventeen had pathologically confirmed final surgical margins of less than 2 mm or focally positive. Brachytherapy catheter placement and treatment delivery were conducted on an outpatient basis. Preplanning was used to determine optimal catheter positions to enhance dose homogeneity of dose delivery. The total HDR boost dose was 15 Gy delivered in 6 fractions of 2.5 Gy over 3 days. Local control, survival, late toxicities (LENT-SOMA), and cosmetic outcome were recorded in follow-up. In addition, factors potentially influencing cosmesis were analyzed by logistic regression analysis. Results: The minimum follow-up is 40 months with a median 50 months. Sixteen patients were alive without disease at last follow-up. There have been no in-breast failures observed. One patient died with brain metastases, and another died of unrelated causes without evidence of disease. Grade 1-2 late toxicities included 39% with hyperpigmentation, 56% with detectable fibrosis, 28% with occasional discomfort, and 11% with visible telangiectasias. Grade 3 toxicity was reported in one patient as persistent discomfort. Sixty-seven percent of patients were considered to have experienced good/excellent cosmetic outcomes. Factors with a direct relationship to adverse cosmetic outcome were extent of surgical defect (p = 0.00001), primary excision volume (p = 0.017), and total excision volume (p = 0.015). Conclusions: For high risk patients who may benefit from increased doses, interstitial HDR

  18. Evaluation of linear array MOSFET detectors for in vivo dosimetry to measure rectal dose in HDR brachytherapy.

    Science.gov (United States)

    Haughey, Aisling; Coalter, George; Mugabe, Koki

    2011-09-01

    The study aimed to assess the suitability of linear array metal oxide semiconductor field effect transistor detectors (MOSFETs) as in vivo dosimeters to measure rectal dose in high dose rate brachytherapy treatments. The MOSFET arrays were calibrated with an Ir192 source and phantom measurements were performed to check agreement with the treatment planning system. The angular dependence, linearity and constancy of the detectors were evaluated. For in vivo measurements two sites were investigated, transperineal needle implants for prostate cancer and Fletcher suites for cervical cancer. The MOSFETs were inserted into the patients' rectum in theatre inside a modified flatus tube. The patients were then CT scanned for treatment planning. Measured rectal doses during treatment were compared with point dose measurements predicted by the TPS. The MOSFETs were found to require individual calibration factors. The calibration was found to drift by approximately 1% ±0.8 per 500 mV accumulated and varies with distance from source due to energy dependence. In vivo results for prostate patients found only 33% of measured doses agreed with the TPS within ±10%. For cervix cases 42% of measured doses agreed with the TPS within ±10%, however of those not agreeing variations of up to 70% were observed. One of the most limiting factors in this study was found to be the inability to prevent the MOSFET moving internally between the time of CT and treatment. Due to the many uncertainties associated with MOSFETs including calibration drift, angular dependence and the inability to know their exact position at the time of treatment, we consider them to be unsuitable for in vivo dosimetry in rectum for HDR brachytherapy.

  19. Vaginal tolerance of CT based image-guided high-dose rate interstitial brachytherapy for gynecological malignancies

    International Nuclear Information System (INIS)

    Murakami, Naoya; Inaba, Koji; Morota, Madoka; Ito, Yoshinori; Itami, Jun; Kasamatsu, Takahiro; Sumi, Minako; Yoshimura, Ryoichi; Harada, Ken; Kitaguchi, Mayuka; Sekii, Shuhei; Takahashi, Kana; Yoshio, Kotaro

    2014-01-01

    Purpose of this study was to identify predictors of vaginal ulcer after CT based three-dimensional image-guided high-dose-rate interstitial brachytherapy (HDR-ISBT) for gynecologic malignancies. Records were reviewed for 44 female (14 with primary disease and 30 with recurrence) with gynecological malignancies treated with HDR-ISBT with or without external beam radiation therapy. The HDR-ISBT applicator insertion was performed with image guidance by trans-rectal ultrasound and CT. The median clinical target volume was 35.5 ml (2.4-142.1 ml) and the median delivered dose in equivalent dose in 2 Gy fractions (EQD 2 ) for target volume D 90 was 67.7 Gy (48.8-94.2 Gy, doses of external-beam radiation therapy and brachytherapy were combined). For re-irradiation patients, median EQD 2 of D 2cc for rectum and bladder, D 0.5cc , D 1cc , D 2cc , D 4cc , D 6cc and D 8cc for vaginal wall was 91.1 Gy, 100.9 Gy, 260.3 Gy, 212.3 Gy, 170.1 Gy, 117.1 Gy, 105.2 Gy, and 94.7 Gy, respectively. For those without prior radiation therapy, median EQD 2 of D 2cc for rectum and bladder, D 0.5cc , D 1cc , D 2cc , D 4cc , D 6cc and D 8cc for vaginal wall was 56.3 Gy, 54.3 Gy, 147.4 Gy, 126.2 Gy, 108.0 Gy, 103.5 Gy, 94.7 Gy, and 80.7 Gy, respectively. Among five patients with vaginal ulcer, three had prior pelvic radiation therapy in their initial treatment and three consequently suffered from fistula formation. On univariate analysis, re-irradiation and vaginal wall D 2cc in EQD 2 was the clinical predictors of vaginal ulcer (p = 0.035 and p = 0.025, respectively). The ROC analysis revealed that vaginal wall D 2cc is the best predictor of vaginal ulcer. The 2-year incidence rates of vaginal ulcer in the patients with vaginal wall D 2cc in EQD 2 equal to or less than 145 Gy and over 145 Gy were 3.7% and 23.5%, respectively, with a statistically significant difference (p = 0.026). Re-irradiation and vaginal D 2cc is a significant predictor of vaginal ulcer after HDR-ISBT for gynecologic

  20. Dosimetric Coverage of the Prostate, Normal Tissue Sparing, and Acute Toxicity with High-Dose-Rate Brachytherapy for Large Prostate Volumes

    Directory of Open Access Journals (Sweden)

    George Yang

    2015-06-01

    Full Text Available ABSTRACTPurposeTo evaluate dosimetric coverage of the prostate, normal tissue sparing, and acute toxicity with HDR brachytherapy for large prostate volumes.Materials and MethodsOne hundred and two prostate cancer patients with prostate volumes >50 mL (range: 5-29 mL were treated with high-dose-rate (HDR brachytherapy ± intensity modulated radiation therapy (IMRT to 4,500 cGy in 25 daily fractions between 2009 and 2013. HDR brachytherapy monotherapy doses consisted of two 1,350-1,400 cGy fractions separated by 2-3 weeks, and HDR brachytherapy boost doses consisted of two 950-1,150 cGy fractions separated by 4 weeks. Twelve of 32 (38% unfavorable intermediate risk, high risk, and very high risk patients received androgen deprivation therapy. Acute toxicity was graded according to the Common Terminology Criteria for Adverse Events (CTCAE version 4.ResultsMedian follow-up was 14 months. Dosimetric goals were achieved in over 90% of cases. Three of 102 (3% patients developed Grade 2 acute proctitis. No variables were significantly associated with Grade 2 acute proctitis. Seventeen of 102 (17% patients developed Grade 2 acute urinary retention. American Urological Association (AUA symptom score was the only variable significantly associated with Grade 2 acute urinary retention (p=0.04. There was no ≥ Grade 3 acute toxicity.ConclusionsDosimetric coverage of the prostate and normal tissue sparing were adequate in patients with prostate volumes >50 mL. Higher pre-treatment AUA symptom scores increased the relative risk of Grade 2 acute urinary retention. However, the overall incidence of acute toxicity was acceptable in patients with large prostate volumes.

  1. Dosimetric coverage of the prostate, normal tissue sparing, and acute toxicity with high-dose-rate brachytherapy for large prostate volumes

    Energy Technology Data Exchange (ETDEWEB)

    Yang, George; Strom, Tobin J.; Shrinath, Kushagra; Mellon, Eric A.; Fernandez, Daniel C.; Biagioli, Matthew C. [Department of Radiation Oncology, H. Lee Moffitt Cancer Center and Research Institute, Tampa, FL (United States); Wilder, Richard B., E-mail: mcbiagioli@yahoo.com [Cancer Treatment Centers of America, Newnan, GA (United States)

    2015-05-15

    Purpose: to evaluate dosimetric coverage of the prostate, normal tissue sparing, and acute toxicity with HDR brachytherapy for large prostate volumes. Materials and methods: one hundred and two prostate cancer patients with prostate volumes >50 mL (range: 5-29 mL) were treated with high-dose-rate (HDR) brachytherapy ± intensity modulated radiation therapy (IMRT) to 4,500 cGy in 25 daily fractions between 2009 and 2013. HDR brachytherapy monotherapy doses consisted of two 1,350-1,400 cGy fractions separated by 2-3 weeks, and HDR brachytherapy boost doses consisted of two 950-1,150 cGy fractions separated by 4 weeks. Twelve of 32 (38%) unfavorable intermediate risk, high risk, and very high risk patients received androgen deprivation therapy. Acute toxicity was graded according to the Common Terminology Criteria for Adverse Events (CTCAE) version 4. Results: median follow-up was 14 months. Dosimetric goals were achieved in over 90% of cases. Three of 102 (3%) patients developed Grade 2 acute proctitis. No variables were significantly associated with Grade 2 acute proctitis. Seventeen of 102 (17%) patients developed Grade 2 acute urinary retention. American Urological Association (AUA) symptom score was the only variable significantly associated with Grade 2 acute urinary retention (p-0.04). There was no ≥ Grade 3 acute toxicity. Conclusions: dosimetric coverage of the prostate and normal tissue sparing were adequate in patients with prostate volumes >50 mL. Higher pre-treatment AUA symptom scores increased the relative risk of Grade 2 acute urinary retention. However, the overall incidence of acute toxicity was acceptable in patients with large prostate volumes. (author)

  2. Sci-Thur PM – Brachytherapy 04: Commissioning and Implementation of a Cobalt-60 High Dose Rate Brachytherapy Source

    Energy Technology Data Exchange (ETDEWEB)

    Dysart, Jonathan [Horizon Health Network (Canada)

    2016-08-15

    An Eckert & Ziegler Bebig Co0.A86 cobalt 60 high dose rate (HDR) brachytherapy source was commissioned for clinical use. Long-lived Co-60 HDR sources offer potential logistical and economic advantages over Ir-192 sources, and should be considered for low to medium workload brachytherapy departments where modest increases in treatment times are not a factor. In optimized plans, the Co-60 source provides a similar dose distribution to Ir-192 despite the difference in radiation energy. By switching to Co-60, source exchange frequency can be reduced by a factor of 20, resulting in overall financial savings of more than 50% compared to Ir-192 sources. In addition, a reduction in Physicist QA workload of roughly 200 hours over the 5 year life of the Co-60 source is also expected. These benefits should be considered against the modest increases in average treatment time compared to those of Ir-192 sources, as well as the centre-specific needs for operating room shielding modification.

  3. Comparison BIPM.RI(I)-K8 of high dose-rate Ir-192 brachytherapy standards for reference air kerma rate of the PTB and the BIPM

    International Nuclear Information System (INIS)

    Kessler, C.; Allisy-Roberts, P.J.; Selbach, H.J.

    2015-01-01

    An indirect comparison of the standards for reference air kerma rate (RAKR) for 192 Ir high dose rate (HDR) brachytherapy sources of the Physikalisch-Technische Bundesanstalt (PTB), Germany, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the PTB in September 2011. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the PTB and the BIPM standards for reference air kerma rate, is 1.0003 with a combined standard uncertainty of 0.0099. (authors)

  4. Calculated organ doses using Monte Carlo simulations in a reference male phantom undergoing HDR brachytherapy applied to localized prostate carcinoma

    International Nuclear Information System (INIS)

    Candela-Juan, Cristian; Perez-Calatayud, Jose; Ballester, Facundo; Rivard, Mark J.

    2013-01-01

    Purpose: The aim of this study was to obtain equivalent doses in radiosensitive organs (aside from the bladder and rectum) when applying high-dose-rate (HDR) brachytherapy to a localized prostate carcinoma using 60 Co or 192 Ir sources. These data are compared with results in a water phantom and with expected values in an infinite water medium. A comparison with reported values from proton therapy and intensity-modulated radiation therapy (IMRT) is also provided. Methods: Monte Carlo simulations in Geant4 were performed using a voxelized phantom described in International Commission on Radiological Protection (ICRP) Publication 110, which reproduces masses and shapes from an adult reference man defined in ICRP Publication 89. Point sources of 60 Co or 192 Ir with photon energy spectra corresponding to those exiting their capsules were placed in the center of the prostate, and equivalent doses per clinical absorbed dose in this target organ were obtained in several radiosensitive organs. Values were corrected to account for clinical circumstances with the source located at various positions with differing dwell times throughout the prostate. This was repeated for a homogeneous water phantom. Results: For the nearest organs considered (bladder, rectum, testes, small intestine, and colon), equivalent doses given by 60 Co source were smaller (8%–19%) than from 192 Ir. However, as the distance increases, the more penetrating gamma rays produced by 60 Co deliver higher organ equivalent doses. The overall result is that effective dose per clinical absorbed dose from a 60 Co source (11.1 mSv/Gy) is lower than from a 192 Ir source (13.2 mSv/Gy). On the other hand, equivalent doses were the same in the tissue and the homogeneous water phantom for those soft tissues closer to the prostate than about 30 cm. As the distance increased, the differences of photoelectric effect in water and soft tissue, and appearance of other materials such as air, bone, or lungs, produced

  5. An experimental MOSFET approach to characterize (192)Ir HDR source anisotropy.

    Science.gov (United States)

    Toye, W C; Das, K R; Todd, S P; Kenny, M B; Franich, R D; Johnston, P N

    2007-09-07

    The dose anisotropy around a (192)Ir HDR source in a water phantom has been measured using MOSFETs as relative dosimeters. In addition, modeling using the EGSnrc code has been performed to provide a complete dose distribution consistent with the MOSFET measurements. Doses around the Nucletron 'classic' (192)Ir HDR source were measured for a range of radial distances from 5 to 30 mm within a 40 x 30 x 30 cm(3) water phantom, using a TN-RD-50 MOSFET dosimetry system with an active area of 0.2 mm by 0.2 mm. For each successive measurement a linear stepper capable of movement in intervals of 0.0125 mm re-positioned the MOSFET at the required radial distance, while a rotational stepper enabled angular displacement of the source at intervals of 0.9 degrees . The source-dosimeter arrangement within the water phantom was modeled using the standardized cylindrical geometry of the DOSRZnrc user code. In general, the measured relative anisotropy at each radial distance from 5 mm to 30 mm is in good agreement with the EGSnrc simulations, benchmark Monte Carlo simulation and TLD measurements where they exist. The experimental approach employing a MOSFET detection system of small size, high spatial resolution and fast read out capability allowed a practical approach to the determination of dose anisotropy around a HDR source.

  6. In vivo assessment of catheter positioning accuracy and prolonged irradiation time on liver tolerance dose after single-fraction 192Ir high-dose-rate brachytherapy

    Directory of Open Access Journals (Sweden)

    Kropf Siegfried

    2011-09-01

    Full Text Available Abstract Background To assess brachytherapy catheter positioning accuracy and to evaluate the effects of prolonged irradiation time on the tolerance dose of normal liver parenchyma following single-fraction irradiation with 192 Ir. Materials and methods Fifty patients with 76 malignant liver tumors treated by computed tomography (CT-guided high-dose-rate brachytherapy (HDR-BT were included in the study. The prescribed radiation dose was delivered by 1 - 11 catheters with exposure times in the range of 844 - 4432 seconds. Magnetic resonance imaging (MRI datasets for assessing irradiation effects on normal liver tissue, edema, and hepatocyte dysfunction, obtained 6 and 12 weeks after HDR-BT, were merged with 3D dosimetry data. The isodose of the treatment plan covering the same volume as the irradiation effect was taken as a surrogate for the liver tissue tolerance dose. Catheter positioning accuracy was assessed by calculating the shift between the 3D center coordinates of the irradiation effect volume and the tolerance dose volume for 38 irradiation effects in 30 patients induced by catheters implanted in nearly parallel arrangement. Effects of prolonged irradiation were assessed in areas where the irradiation effect volume and tolerance dose volume did not overlap (mismatch areas by using a catheter contribution index. This index was calculated for 48 irradiation effects induced by at least two catheters in 44 patients. Results Positioning accuracy of the brachytherapy catheters was 5-6 mm. The orthogonal and axial shifts between the center coordinates of the irradiation effect volume and the tolerance dose volume in relation to the direction vector of catheter implantation were highly correlated and in first approximation identically in the T1-w and T2-w MRI sequences (p = 0.003 and p p = 0.001 and p = 0.004, respectively. There was a significant shift of the irradiation effect towards the catheter entry site compared with the planned dose

  7. Intraoperative HDR implant boost for breast cancer (preliminary results)

    International Nuclear Information System (INIS)

    Rodriguez, I.; Torre, M. de la; Gonzalez, E.; Bourel, V.

    1996-01-01

    Introduction: In spite of the fact that it is been discussed whether or not a boost is necessary for all conservative treated breast cancer patients, it is a generalized radiotherapy practice. Since september 1993 we developed a breast conservative protocol for early stage breast cancer (T1-T2) with intraoperative HDR implant boost. Side effects, cosmetic results and recurrence rates are reviewed. Method and Material: From September 1993 we treated 55 patients with intraoperative HDR implant boost to the lumpectomy site for clinical T1 or T2 invasive breast cancer, followed by external megavoltage radiotherapy to the entire breast. We used the Nucletron microselectron HDR remote afterloading system with flexible implant tubes. The geometric distribution of the tubes was performed according to the 'Paris' configuration. Each implant was evaluated by calculating the dose-volume natural histograms. The HDR fractionation schedule consists of three fractions of 4.5Gy each given at least 48 hs apart, and starting between 48-72hs from surgical procedure. The external radiotherapy to the entire breast started one week after the completion of brachytherapy, using conventional fractionation of 5 fractions per week, 1,8Gy per fraction up to 45-50Gy. Results: So far there is not any local recurrence, but medium follow up is only 18 months. We did not observe any acute damage and the cosmetic outcome was 60% excellent, 30% good and 10% acceptable. Two patients developed localized fibrosis, in both the implant involved the submamary fold. Conclusion: The intraoperative implant is the most accurate way to localize the lumpectomy site, to define the target volume, decrease the total treatment time and avoid a second anesthetic procedure without delaying the inpatient time or the initial wound healing process

  8. SU-F-T-32: Evaluation of the Performance of a Multiple-Array-Diode Detector for Quality Assurance Tests in High-Dose-Rate Brachytherapy with Ir-192 Source

    Energy Technology Data Exchange (ETDEWEB)

    Harpool, K; De La Fuente Herman, T; Ahmad, S; Ali, I [University of Oklahoma Health Sciences Center, Oklahoma City, OK (United States)

    2016-06-15

    Purpose: To evaluate the performance of a two-dimensional (2D) array-diode- detector for geometric and dosimetric quality assurance (QA) tests of high-dose-rate (HDR) brachytherapy with an Ir-192-source. Methods: A phantom setup was designed that encapsulated a two-dimensional (2D) array-diode-detector (MapCheck2) and a catheter for the HDR brachytherapy Ir-192 source. This setup was used to perform both geometric and dosimetric quality assurance for the HDR-Ir192 source. The geometric tests included: (a) measurement of the position of the source and (b) spacing between different dwell positions. The dosimteric tests include: (a) linearity of output with time, (b) end effect and (c) relative dose verification. The 2D-dose distribution measured with MapCheck2 was used to perform the previous tests. The results of MapCheck2 were compared with the corresponding quality assurance testes performed with Gafchromic-film and well-ionization-chamber. Results: The position of the source and the spacing between different dwell-positions were reproducible within 1 mm accuracy by measuring the position of maximal dose using MapCheck2 in contrast to the film which showed a blurred image of the dwell positions due to limited film sensitivity to irradiation. The linearity of the dose with dwell times measured from MapCheck2 was superior to the linearity measured with ionization chamber due to higher signal-to-noise ratio of the diode readings. MapCheck2 provided more accurate measurement of the end effect with uncertainty < 1.5% in comparison with the ionization chamber uncertainty of 3%. Although MapCheck2 did not provide absolute calibration dosimeter for the activity of the source, it provided accurate tool for relative dose verification in HDR-brachytherapy. Conclusion: The 2D-array-diode-detector provides a practical, compact and accurate tool to perform quality assurance for HDR-brachytherapy with an Ir-192 source. The diodes in MapCheck2 have high radiation sensitivity and

  9. Pulsed dose rate brachytherapy – is it the right way?

    Directory of Open Access Journals (Sweden)

    Janusz Skowronek

    2010-10-01

    Full Text Available Pulsed dose rate (PDR-BT treatment is a brachytherapy modality that combines physical advantages of high-doserate (HDR-BT technology (isodose optimization, radiation safety with the radiobiological advantages of low-dose-rate (LDR-BT brachytherapy. Pulsed brachytherapy consists of using stronger radiation source than for LDR-BT and producing series of short exposures of 10 to 30 minutes in every hour to approximately the same total dose in the sameoverall time as with the LDR-BT. Modern afterloading equipment offers certain advantages over interstitial or intracavitaryinsertion of separate needles, tubes, seeds or wires. Isodose volumes in tissues can be created flexibly by a combinationof careful placement of the catheter and the adjustment of the dwell times of the computerized stepping source.Automatic removal of the radiation sources into a shielded safe eliminates radiation exposures to staff and visitors.Radiation exposure is also eliminated to the staff who formerly loaded and unloaded multiplicity of radioactive sources into the catheters, ovoids, tubes etc. This review based on summarized clinical investigations, analyses the feasibility and the background to introduce this brachytherapy technique and chosen clinical applications of PDR-BT.

  10. The IPEM code of practice for determination of the reference air kerma rate for HDR 192Ir brachytherapy sources based on the NPL air kerma standard

    International Nuclear Information System (INIS)

    Bidmead, A M; Sander, T; Nutbrown, R F; Locks, S M; Lee, C D; Aird, E G A; Flynn, A

    2010-01-01

    This paper contains the recommendations of the high dose rate (HDR) brachytherapy working party of the UK Institute of Physics and Engineering in Medicine (IPEM). The recommendations consist of a Code of Practice (COP) for the UK for measuring the reference air kerma rate (RAKR) of HDR 192 Ir brachytherapy sources. In 2004, the National Physical Laboratory (NPL) commissioned a primary standard for the realization of RAKR of HDR 192 Ir brachytherapy sources. This has meant that it is now possible to calibrate ionization chambers directly traceable to an air kerma standard using an 192 Ir source (Sander and Nutbrown 2006 NPL Report DQL-RD 004 (Teddington: NPL) http://publications.npl.co.uk). In order to use the source specification in terms of either RAKR, .K R (ICRU 1985 ICRU Report No 38 (Washington, DC: ICRU); ICRU 1997 ICRU Report No 58 (Bethesda, MD: ICRU)), or air kerma strength, S K (Nath et al 1995 Med. Phys. 22 209-34), it has been necessary to develop algorithms that can calculate the dose at any point around brachytherapy sources within the patient tissues. The AAPM TG-43 protocol (Nath et al 1995 Med. Phys. 22 209-34) and the 2004 update TG-43U1 (Rivard et al 2004 Med. Phys. 31 633-74) have been developed more fully than any other protocol and are widely used in commercial treatment planning systems. Since the TG-43 formalism uses the quantity air kerma strength, whereas this COP uses RAKR, a unit conversion from RAKR to air kerma strength was included in the appendix to this COP. It is recommended that the measured RAKR determined with a calibrated well chamber traceable to the NPL 192 Ir primary standard is used in the treatment planning system. The measurement uncertainty in the source calibration based on the system described in this COP has been reduced considerably compared to other methods based on interpolation techniques.

  11. High-dose rate brachytherapy in the treatment of prostate cancer: acute toxicity and biochemical behavior analysis

    International Nuclear Information System (INIS)

    Esteves, Sergio Carlos Barros; Oliveira, Antonio Carlos Zuliani de; Cardoso, Herbeni; Tagawa, Eduardo Komai; Castelo, Roberto; D'Imperio, Marcio

    2006-01-01

    Objective: this study focuses on the biochemical response of the following variables: prostate volume, prostate-specific antigen (PSA) value, Gleason scores, staging, the risk of the disease, and hormone therapy. Objective: in the period between February of 1998 and July of 2001, 46 patients with prostate cancer were treated with radiotherapy, in a combination of teletherapy and high-dose rate (HDR) brachytherapy. The age ranged from 51 to 79 years (averaging 66.4 years). T1c stage was the most frequent one: 30 (65%). The Gleason score was below 7 in 78% of the patients. PSA ranged from 3.4 to 33.3, being below 10 in 39% of the cases. The average prostatic volume was 32.3 cc. Twenty-eight percent of the patients received hormone therapy. Teletherapy dose ranged from 45 to 50.4 Gy, associated to four fractions of 4 Gy of HDR brachytherapy. Results: the follow-up period varied from 6 to 43 months. Four patients missed the follow-up and four died (one due to the disease). Out of the 39 patients that were analyzed, 76% presented a less than 1.5 PSA. None of the analyzed variables were found to be of statistical significance (p > 0.05) regarding biochemical control. Conclusion: the use of HDR brachytherapy was found to be effective in the treatment of prostate cancer and, in this study, the variables considered as prognostic factors did not interfere in the biochemical control. (author)

  12. HDR and LDR Brachytherapy in the Treatment of Lip Cancer: the Experience of the Catalan Institute of Oncology.

    Science.gov (United States)

    Ayerra, Arrate Querejeta; Mena, Estefanía Palacios; Fabregas, Joan Pera; Miguelez, Cristina Gutiérrez; Guedea, Ferran

    2010-03-01

    Lip cancer can be treated by surgery, external radiotherapy, and/or brachytherapy (BT). In recent years, BT has become increasingly favored for this type of cancer. The aim of the present study was to analyze local control and survival of patients treated at our institution between July 1989 and June 2008. We performed a retrospective study of 121 patients (109 males and 12 females) who underwent lip cancer brachytherapy from July 1989 to June 2008. Median age was 67 years and median follow-up was 31.8 months (range 20-188 months). Out of 121 patients, 100 (82.6%) were treated with low dose rate (LDR) BT while the remaining 21 patients (17.4%) received high dose rate (HDR) BT. The most common cell type was squamous cell carcinoma (115 cases; 95%) and most tumors were located on the lower lip (107 patients; 88.4%). Most cases were either stage T1 (62 patients; 51.2%), or T2 (44 cases; 36.4%). After 15 years of follow-up, overall survival was 89.5%, cause-specific survival 97.8%, and disease-free survival 86.6%. Local, regional, and distant control at 15 years were 90%, 92%, and 98.8%, respectively. Grade 3 mucosal toxicity was observed in 23% of patients treated with LDR compared to 33% of HDR patients, and grade 4 mucosal toxicity in 9% versus 0% in the HDR group. Our findings confirm that brachytherapy is an effective treatment for lip cancer. The results from our series are in line with those published elsewhere. Based on our limited data, HDR appears to be equally as good as LDR, although this needs to be confirmed by further studies.

  13. Brachytherapy source calibration, reviews, and consistency of 192Ir high-dose rate afterloading sources supplied over the period of 10 years: a retrospective analysis

    International Nuclear Information System (INIS)

    Nagappan, Balasubramanian; Kumar, Yogesh; Patel, Narayan P.; Dhull, Anil Kumar; Kaushal, Vivek

    2015-01-01

    Measurement and verification of strength of monomodal high-dose rate (mHDR) 192 Ir source supplied by the vendor is a major part of quality assurance program. Reference air kerma rate (RAKR) or air kerma strength (AKS) is the recommended quantity to specify the strength of gamma emitting brachytherapy sources. Physicist in our institution performed the source calibration as soon as each 192 Ir new source was loaded on the mHDR afterloading machine. The AKS accurately measured using a physikalisch technische werkstatten (PTW) re-entrant chamber-electrometer system in a scatter-free geometry was used to compute the air kerma rate (AKR) at one-meter distance in the air. To ensure accurate dose delivery to brachytherapy patients, measured AKS or RAKR should be entered correctly in both HDR treatment console station (TCS) as well as treatment planning system (TPS) associated with it. The clinical outcome mainly depends not only on the accuracy of the source strength measurement in the hospital but also on the correct source strength entered into both TCS and TPS software. A retrospective study on 22 mHDR V2 sources supplied by the vendor for the period of 10 years was taken up to access the accuracy of source strength supplied to the Radiotherapy department. The results are analyzed and reported. The accuracy in measured RAKR of all 22 sources supplied by vendor was well within the tolerance limits set by the national regulatory body and international recommendations. The deviations observed between measured RAKR versus manufacturer's quoted RAKR were in the range from -1.71% to +1.15%. In conclusion, the measured RAKR have good agreement with vendor quoted RAKR values. (author)

  14. Dosimetric study of surface applicators of HDR brachytherapy GammaMed Plus equipment

    International Nuclear Information System (INIS)

    Reyes-Rivera, E.; Sosa, M.; Reyes, U.; Jesús Bernal-Alvarado, José de; Córdova, T.; Gil-Villegas, A.; Monzón, E.

    2014-01-01

    The cone type surface applicators used in HDR brachytherapy for treatment of small skin lesions are an alternative to be used with both electron beams and orthovoltage X-ray equipment. For a good treatment planning is necessary to know the dose distribution of these applicators, which can be obtained by experimental measurement and Monte Carlo simulation as well. In this study the dose distribution of surface applicators of 3 and 3.5 cm diameter, respectively of HDR brachytherapy GammaMed Plus equipment has been estimated using the Monte Carlo method, MCNP code. The applicators simulated were placed on the surface of a water phantom of 20 × 20 × 20 cm and the dose was calculated at depths from 0 to 3 cm with increments of 0.25 mm. The dose profiles obtained at depth show the expected gradients for surface therapy

  15. Dosimetric study of surface applicators of HDR brachytherapy GammaMed Plus equipment

    Energy Technology Data Exchange (ETDEWEB)

    Reyes-Rivera, E., E-mail: eric-1985@fisica.ugto.mx, E-mail: modesto@fisica.ugto.mx, E-mail: uvaldoreyes@fisica.ugto.mx; Sosa, M., E-mail: eric-1985@fisica.ugto.mx, E-mail: modesto@fisica.ugto.mx, E-mail: uvaldoreyes@fisica.ugto.mx; Reyes, U., E-mail: eric-1985@fisica.ugto.mx, E-mail: modesto@fisica.ugto.mx, E-mail: uvaldoreyes@fisica.ugto.mx; Jesús Bernal-Alvarado, José de, E-mail: bernal@fisica.ugto.mx, E-mail: theo@fisica.ugto.mx, E-mail: gil@fisica.ugto.mx; Córdova, T., E-mail: bernal@fisica.ugto.mx, E-mail: theo@fisica.ugto.mx, E-mail: gil@fisica.ugto.mx; Gil-Villegas, A., E-mail: bernal@fisica.ugto.mx, E-mail: theo@fisica.ugto.mx, E-mail: gil@fisica.ugto.mx [División de Ciencias e Ingenierías, Universidad de Guanajuato, 37150 León, Gto. (Mexico); Monzón, E., E-mail: emonzon@imss.gob.mx [Unidad de Alta Especialidad No.1, Instituto Mexicano del Seguro Social, Léon, Gto. (Mexico)

    2014-11-07

    The cone type surface applicators used in HDR brachytherapy for treatment of small skin lesions are an alternative to be used with both electron beams and orthovoltage X-ray equipment. For a good treatment planning is necessary to know the dose distribution of these applicators, which can be obtained by experimental measurement and Monte Carlo simulation as well. In this study the dose distribution of surface applicators of 3 and 3.5 cm diameter, respectively of HDR brachytherapy GammaMed Plus equipment has been estimated using the Monte Carlo method, MCNP code. The applicators simulated were placed on the surface of a water phantom of 20 × 20 × 20 cm and the dose was calculated at depths from 0 to 3 cm with increments of 0.25 mm. The dose profiles obtained at depth show the expected gradients for surface therapy.

  16. Interstitial high-dose-rate brachytherapy in the treatment of base of tongue carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Takacsi-Nagy, Z.; Polgar, C.; Somogyi, A.; Major, T.; Fodor, J.; Nemeth, G. [Dept. of Radiotherapy, National Inst. of Oncology, Budapest (Hungary); Oberna, F. [Dept. of Maxillofacial Surgery, St. Rokus Hospital, Budapest (Hungary); Remenar, E.; Kasler, M. [Dept. of Head and Neck, Maxillofacial and Reconstructive Plastic Surgery, National Inst. of Oncology, Budapest (Hungary)

    2004-12-01

    Background and purpose: to date none of the studies examined the feasibility and efficacy of interstitial high-dose-rate (HDR) brachytherapy in the treatment of carcinoma of the tongue base. Therefore the aim of this study was to contribute to this issue. Patients and methods: between 1992 and 2000 37 patients (mean age 55 years) with T1-4 and NO-3 carcinoma of the base of tongue were presented. Neck dissection was carried out in twelve cases (32%). 30 patients with advanced stage received brachytherapy boost after 50-66.5 Gy (mean, 60 Gy) locoregional external beam irradiation (EBI) and 7 patients with early stage (T1-2, NO) were managed locally with wide tumor excision and sole brachytherapy. 4 of them underwent neck dissection and the others were subjected to 50 Gy regional EBI. The mean dose of boost and sole brachytherapy was 18 Gy and 28 Gy, respectively. Results: the median follow-up time for surviving patients was 51 months. The 7 sole brachytherapy patients are living with no evidence of disease. For patients treated with EBI and brachytherapy boost, the 5-year actuarial rate of local, locoregional recurrence-free and overall survival was 60%, 52% and 46%, respectively. For all patients in univariate analysis larger tumor size (T4 vs. T1-3) was significant negative predictor of local (RR: 7.23) and locoregional control (RR: 3.87), but nodal involvement was not. Delayed soft tissue ulceration and osteoradionecrosis occurred in 4 (13%) EBI and brachytherapy treated patients. None of the sole brachytherapy patients experienced severe late radiation toxicity. Conclusion: EBI combined with interstitial HDR brachytherapy boost result in acceptable local tumor control with low incidence of late side effects in patients with advanced disease. Fractionated sole HDR brachytherapy following tumor excision is a feasible treatment option for patients with early stage cancer and gives excellent local results. (orig.)

  17. Palliative interstitial HDR brachytherapy for recurrent rectal cancer. Implantation techniques and results

    International Nuclear Information System (INIS)

    Kolotas, C.; Roeddiger, S.; Martin, T.; Tselis, N.; Baltas, D.; Zamboglou, N.; Strassmann, G.; Aebersold, D.M.

    2003-01-01

    Purpose: To report the methods and clinical results of CT-based interstitial high-dose-rate (HDR) brachytherapy procedures for the palliative treatment of recurrent rectal cancer. Patients and Methods: A total of 44 brachytherapy implants were performed in 38 patients. CT-guided catheter implants were performed in 34 patients under local anesthesia and sedation, and four patients were implanted intraoperatively. Of 40 CT-guided implants, 20 were done using metallic needles introduced via the sacrum and 20 were transperineal implants of plastic tubes in the presacral region. Postimplant CT scans were used for three-dimensional (3-D) conformal brachytherapy planning. Patients implanted with metallic needles were given a single fraction of 10-15 Gy using HDR 192 Ir, and those who received transperineal implants of plastic catheters were given fractionated brachytherapy, 5 Gy twice daily to a total dose of 30-40 Gy. The median tumor volume was 225 cm 3 with a range of 41-2,103 cm 3 . Results: After a median follow-up of 23.4 months, a total of 13/38 patients were alive. The median postbrachytherapy survival was 15 months with 18 of the 25 deaths due to distant metastases. Tumor response was as follows: 6/38 partial remission, 28/38 stable disease, and 4/38 local progression. A planning target volume (PTV) coverage > 85% was achieved in 42/44 implants. The treatment was well tolerated, and no acute complications were observed. One patient developed a fistula after 8 months. Pain relief was recorded in 34 patients (89.5%), and the median duration of this palliative effect was 5 months with a range of 1-13 months. Conclusions: Interstitial HDR brachytherapy is a valuable tool for the delivery of high doses and achieves effective palliation in recurrent rectal carcinoma. (orig.)

  18. Palliative interstitial HDR brachytherapy for recurrent rectal cancer. Implantation techniques and results

    Energy Technology Data Exchange (ETDEWEB)

    Kolotas, C. [Dept. of Radiation Oncology, Offenbach Hospital, Offenbach (Germany); Dept. of Radio-Oncology, Univ. of Bern, Inselspital, Bern (Switzerland); Roeddiger, S.; Martin, T.; Tselis, N.; Baltas, D.; Zamboglou, N. [Dept. of Radiation Oncology, Offenbach Hospital, Offenbach (Germany); Strassmann, G. [Dept. of Radiotherapy, Univ. Hospital, Philipps Univ., Marburg (Germany); Aebersold, D.M. [Dept. of Radio-Oncology, Univ. of Bern, Inselspital, Bern (Switzerland)

    2003-07-01

    Purpose: To report the methods and clinical results of CT-based interstitial high-dose-rate (HDR) brachytherapy procedures for the palliative treatment of recurrent rectal cancer. Patients and Methods: A total of 44 brachytherapy implants were performed in 38 patients. CT-guided catheter implants were performed in 34 patients under local anesthesia and sedation, and four patients were implanted intraoperatively. Of 40 CT-guided implants, 20 were done using metallic needles introduced via the sacrum and 20 were transperineal implants of plastic tubes in the presacral region. Postimplant CT scans were used for three-dimensional (3-D) conformal brachytherapy planning. Patients implanted with metallic needles were given a single fraction of 10-15 Gy using HDR {sup 192}Ir, and those who received transperineal implants of plastic catheters were given fractionated brachytherapy, 5 Gy twice daily to a total dose of 30-40 Gy. The median tumor volume was 225 cm{sup 3} with a range of 41-2,103 cm{sup 3}. Results: After a median follow-up of 23.4 months, a total of 13/38 patients were alive. The median postbrachytherapy survival was 15 months with 18 of the 25 deaths due to distant metastases. Tumor response was as follows: 6/38 partial remission, 28/38 stable disease, and 4/38 local progression. A planning target volume (PTV) coverage > 85% was achieved in 42/44 implants. The treatment was well tolerated, and no acute complications were observed. One patient developed a fistula after 8 months. Pain relief was recorded in 34 patients (89.5%), and the median duration of this palliative effect was 5 months with a range of 1-13 months. Conclusions: Interstitial HDR brachytherapy is a valuable tool for the delivery of high doses and achieves effective palliation in recurrent rectal carcinoma. (orig.)

  19. A Novel Form of Breast Intraoperative Radiation Therapy With CT-Guided High-Dose-Rate Brachytherapy: Results of a Prospective Phase 1 Clinical Trial

    International Nuclear Information System (INIS)

    Showalter, Shayna L.; Petroni, Gina; Trifiletti, Daniel M.; Libby, Bruce; Schroen, Anneke T.; Brenin, David R.; Dalal, Parchayi; Smolkin, Mark; Reardon, Kelli A.; Showalter, Timothy N.

    2016-01-01

    Purpose: Existing intraoperative radiation therapy (IORT) techniques are criticized for the lack of image guided treatment planning and energy deposition with, at times, poor resultant dosimetry and low radiation dose. We pioneered a novel method of IORT that incorporates customized, computed tomography (CT)-based treatment planning and high-dose-rate (HDR) brachytherapy to overcome these drawbacks: CT-HDR-IORT. Methods and Materials: A phase 1 study was conducted to demonstrate the feasibility and safety of CT-HDR-IORT. Eligibility criteria included age ≥50 years, invasive or in situ breast cancer, tumor size <3 cm, and N0 disease. Patients were eligible before or within 30 days of breast-conserving surgery (BCS). BCS was performed, and a multilumen balloon catheter was placed. CT images were obtained, a customized HDR brachytherapy plan was created, and a dose of 12.5 Gy was delivered to 1-cm depth from the balloon surface. The catheter was removed, and the skin was closed. The primary endpoints were feasibility and acute toxicity. Feasibility was defined as IORT treatment interval (time from CT acquisition until IORT completion) ≤90 minutes. The secondary endpoints included dosimetry, cosmetic outcome, quality of life, and late toxicity. Results: Twenty-eight patients were enrolled. The 6-month follow-up assessments were completed by 93% of enrollees. The median IORT treatment interval was 67.2 minutes (range, 50-108 minutes). The treatment met feasibility criteria in 26 women (93%). The dosimetric goals were met in 22 patients (79%). There were no Radiation Therapy Oncology Group grade 3+ toxicities; 6 patients (21%) experienced grade 2 events. Most patients (93%) had good/excellent cosmetic outcomes at the last follow-up visit. Conclusions: CT-HDR-IORT is feasible and safe. This promising approach for a conformal, image-based, higher-dose breast IORT is being evaluated in a phase 2 trial.

  20. A Novel Form of Breast Intraoperative Radiation Therapy With CT-Guided High-Dose-Rate Brachytherapy: Results of a Prospective Phase 1 Clinical Trial

    Energy Technology Data Exchange (ETDEWEB)

    Showalter, Shayna L., E-mail: snl2t@virginia.edu [Division of Surgical Oncology, Department of Surgery, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Petroni, Gina [Division of Translation Research and Applied Statistics, Department of Public Health Sciences, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Trifiletti, Daniel M.; Libby, Bruce [Department of Radiation Oncology, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Schroen, Anneke T.; Brenin, David R. [Division of Surgical Oncology, Department of Surgery, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Dalal, Parchayi [Department of Radiation Oncology, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Smolkin, Mark [Division of Translation Research and Applied Statistics, Department of Public Health Sciences, University of Virginia School of Medicine, Charlottesville, Virginia (United States); Reardon, Kelli A.; Showalter, Timothy N. [Department of Radiation Oncology, University of Virginia School of Medicine, Charlottesville, Virginia (United States)

    2016-09-01

    Purpose: Existing intraoperative radiation therapy (IORT) techniques are criticized for the lack of image guided treatment planning and energy deposition with, at times, poor resultant dosimetry and low radiation dose. We pioneered a novel method of IORT that incorporates customized, computed tomography (CT)-based treatment planning and high-dose-rate (HDR) brachytherapy to overcome these drawbacks: CT-HDR-IORT. Methods and Materials: A phase 1 study was conducted to demonstrate the feasibility and safety of CT-HDR-IORT. Eligibility criteria included age ≥50 years, invasive or in situ breast cancer, tumor size <3 cm, and N0 disease. Patients were eligible before or within 30 days of breast-conserving surgery (BCS). BCS was performed, and a multilumen balloon catheter was placed. CT images were obtained, a customized HDR brachytherapy plan was created, and a dose of 12.5 Gy was delivered to 1-cm depth from the balloon surface. The catheter was removed, and the skin was closed. The primary endpoints were feasibility and acute toxicity. Feasibility was defined as IORT treatment interval (time from CT acquisition until IORT completion) ≤90 minutes. The secondary endpoints included dosimetry, cosmetic outcome, quality of life, and late toxicity. Results: Twenty-eight patients were enrolled. The 6-month follow-up assessments were completed by 93% of enrollees. The median IORT treatment interval was 67.2 minutes (range, 50-108 minutes). The treatment met feasibility criteria in 26 women (93%). The dosimetric goals were met in 22 patients (79%). There were no Radiation Therapy Oncology Group grade 3+ toxicities; 6 patients (21%) experienced grade 2 events. Most patients (93%) had good/excellent cosmetic outcomes at the last follow-up visit. Conclusions: CT-HDR-IORT is feasible and safe. This promising approach for a conformal, image-based, higher-dose breast IORT is being evaluated in a phase 2 trial.

  1. A study of optimization techniques in HDR brachytherapy for the prostate

    Science.gov (United States)

    Pokharel, Ghana Shyam

    Several studies carried out thus far are in favor of dose escalation to the prostate gland to have better local control of the disease. But optimal way of delivery of higher doses of radiation therapy to the prostate without hurting neighboring critical structures is still debatable. In this study, we proposed that real time high dose rate (HDR) brachytherapy with highly efficient and effective optimization could be an alternative means of precise delivery of such higher doses. This approach of delivery eliminates the critical issues such as treatment setup uncertainties and target localization as in external beam radiation therapy. Likewise, dosimetry in HDR brachytherapy is not influenced by organ edema and potential source migration as in permanent interstitial implants. Moreover, the recent report of radiobiological parameters further strengthen the argument of using hypofractionated HDR brachytherapy for the management of prostate cancer. Firstly, we studied the essential features and requirements of real time HDR brachytherapy treatment planning system. Automating catheter reconstruction with fast editing tools, fast yet accurate dose engine, robust and fast optimization and evaluation engine are some of the essential requirements for such procedures. Moreover, in most of the cases we performed, treatment plan optimization took significant amount of time of overall procedure. So, making treatment plan optimization automatic or semi-automatic with sufficient speed and accuracy was the goal of the remaining part of the project. Secondly, we studied the role of optimization function and constraints in overall quality of optimized plan. We have studied the gradient based deterministic algorithm with dose volume histogram (DVH) and more conventional variance based objective functions for optimization. In this optimization strategy, the relative weight of particular objective in aggregate objective function signifies its importance with respect to other objectives

  2. High-dose-rate afterloading brachytherapy in carcinoma of the cervix: an experience of 1992 patients

    International Nuclear Information System (INIS)

    Lorvidhaya, Vicharn; Tonusin, Anun; Changwiwit, Witit; Chitapanarux, Imjai; Srisomboon, Jatupol; Wanwilairat, Somsak; Chawapun, Nisa; Sukthomya, Vimol

    2000-01-01

    Purpose: To report the results of radiation therapy in carcinoma of the cervix treated by external irradiation and high-dose-rate (HDR) intracavitary brachytherapy. Methods and Materials: This is a retrospective analysis of 2063 patients with histologically proven carcinoma of the cervix treated by external irradiation and HDR intracavitary brachytherapy between March 1985-December 1991. The Kaplan-Meier method was used for survival and disease-free survival analysis. Late complications in the bowel and bladder were calculated actuarially. Results: There were 71 patients who did not complete the course of irradiation so only 1992 patients were retrospectively analyzed for survival. There were 2 patients (0.1%) in Stage IA, 211 (10.2%) Stage IB, 225 (10.9%) in Stage IIA, 902 (43.7%) in Stage IIB, 14 (0.7%) in Stage IIIA, 675 (32.7%) in Stage IIIB, 16 (0.8%) in Stage IVA, and 16 (0.8%) in Stage IVB. The median follow-up time was 96 months. The actuarial 5-year disease-free survival rate was 79.5%, 70.0%, 59.4%, 46.1%, 32.3%, 7.8%, and 23.1% for Stage IB, IIA, IIB, IIIA, IIIB, IVA, and IVB respectively. The actuarial 5-year disease-free survival rate for Stage IB 1 and IB 2 squamous cell carcinoma was 88.7% and 67.0%. The actuarial 5-year overall survival rate was 86.3%, 81.1%, 73.0%, 50.3%, 47.8%, 7.8%, and 30.8% for Stage IB, IIA, IIB, IIIA, IIIB, IVA, and IVB respectively. Pattern of failure revealed 20.8% local recurrence, 18.7% distant metastases, and 4% in both. The late complication rate Grade 3 and 4 (RTOG) for bowel and bladder combined was 7.0% with 1.9% Grade 4. Conclusion: HDR brachytherapy used in this series produced pelvic control and survival rates comparable to other LDR series

  3. Anatomy-based inverse optimization in high-dose-rate brachytherapy combined with hypofractionated external beam radiotherapy for localized prostate cancer: Comparison of incidence of acute genitourinary toxicity between anatomy-based inverse optimization and geometric optimization

    International Nuclear Information System (INIS)

    Akimoto, Tetsuo; Katoh, Hiroyuki; Kitamoto, Yoshizumi; Shirai, Katsuyuki; Shioya, Mariko; Nakano, Takashi

    2006-01-01

    Purpose: To evaluate the advantages of anatomy-based inverse optimization (IO) in planning high-dose-rate (HDR) brachytherapy. Methods and Materials: A total of 114 patients who received HDR brachytherapy (9 Gy in two fractions) combined with hypofractionated external beam radiotherapy (EBRT) were analyzed. The dose distributions of HDR brachytherapy were optimized using geometric optimization (GO) in 70 patients and by anatomy-based IO in the remaining 44 patients. The correlation between the dose-volume histogram parameters, including the urethral dose and the incidence of acute genitourinary (GU) toxicity, was evaluated. Results: The averaged values of the percentage of volume receiving 80-150% of the prescribed minimal peripheral dose (V 8 -V 15 ) of the urethra generated by anatomy-based IO were significantly lower than the corresponding values generated by GO. Similarly, the averaged values of the minimal dose received by 5-50% of the target volume (D 5 -D 5 ) obtained using anatomy-based IO were significantly lower than those obtained using GO. Regarding acute toxicity, Grade 2 or worse acute GU toxicity developed in 23% of all patients, but was significantly lower in patients for whom anatomy-based IO (16%) was used than in those for whom GO was used (37%), consistent with the reduced urethral dose (p <0.01). Conclusion: The results of this study suggest that anatomy-based IO is superior to GO for dose optimization in HDR brachytherapy for prostate cancer

  4. Preliminary results of concurrent chemotherapy and radiation therapy using high-dose-rate brachytherapy for cervical cancer

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Kyung Ja; Lee, Ji Hye; Lee, Re Na; Suh, Hyun Suk [Ewha Womans University College of Medicine, Seoul (Korea, Republic of)

    2006-09-15

    To determine the efficacy and safety of concurrent chemotherapy and radiation therapy with high-dose-rate brachytherapy for cervical cancer. From January 2001 to December 2002, 30 patients with cervical cancer were treated with concurrent chemotherapy (cisplatin and 5-FU) and definitive radiation therapy. The median age was 58 (range 34 {approx} 74) year old. The pathology of the biopsy sections was squamous cell carcinoma in 29 patients and one was adenocarcinoma. The distribution to FIGO staging system was as follow: stage IB, 7 (23%); IIA, 3 (10%); IIB, 12 (40%); IIIA, 3 (10%); IIIB, 5 (17%). All patients received pelvic external beam irradiation (EBRT) to a total dose of 45 {approx} 50.4 Gy (median: 50.4 Gy) over 5 {approx} 5.5 weeks. Ir-192 HDR intracavity brachytherapy (ICBT) was given after a total dose of 41.1 Gy. HDR-ICBT was performed twice a week, with a fraction point. A dose of 4 Gy and median dose to point A was 28 Gy (range: 16 {approx} 32 Gy) in 7 fractions. The median cumulative biologic effective dose (BED) at point A (EBRT + ICBT) was 88 Gy{sub 10} (range:77 {approx} 94 Gy{sub 10}). The median cumulative BED at ICRU 38 reference point (EBRT + ICBT) was 131 Gy{sub 3} (range: 122 {approx} 140 Gy{sub 3}) at point A, 109 Gy{sub 3} (range:88{approx} 125 Gy{sub 3}) at the rectum and 111 Gy{sub 3} (range: 91 {approx} 123 Gy{sub 3}) at the urinary bladder. Cisplatin (60 mg/m{sup 2}) and 5-FU (1,000 mg/m{sup 2}) was administered intravenously at 2 weeks interval from the first day of radiation for median 5 (range:2 {approx} 6) cycles. The assessment was performed at 1 month after completion of radiation therapy by clinical examination and CT scan. The median follow-up time was 36 months (range:8{approx} 50 months). The complete response rate after concurrent chemo radiation therapy was 93.3%. The 3-yr actuarial pelvic control rate was 87% and 3-yr actuarial overall survival and disease-free survival rate was 93% and 87%, respectively. The local failure

  5. Preliminary results of concurrent chemotherapy and radiation therapy using high-dose-rate brachytherapy for cervical cancer

    International Nuclear Information System (INIS)

    Lee, Kyung Ja; Lee, Ji Hye; Lee, Re Na; Suh, Hyun Suk

    2006-01-01

    To determine the efficacy and safety of concurrent chemotherapy and radiation therapy with high-dose-rate brachytherapy for cervical cancer. From January 2001 to December 2002, 30 patients with cervical cancer were treated with concurrent chemotherapy (cisplatin and 5-FU) and definitive radiation therapy. The median age was 58 (range 34 ∼ 74) year old. The pathology of the biopsy sections was squamous cell carcinoma in 29 patients and one was adenocarcinoma. The distribution to FIGO staging system was as follow: stage IB, 7 (23%); IIA, 3 (10%); IIB, 12 (40%); IIIA, 3 (10%); IIIB, 5 (17%). All patients received pelvic external beam irradiation (EBRT) to a total dose of 45 ∼ 50.4 Gy (median: 50.4 Gy) over 5 ∼ 5.5 weeks. Ir-192 HDR intracavity brachytherapy (ICBT) was given after a total dose of 41.1 Gy. HDR-ICBT was performed twice a week, with a fraction point. A dose of 4 Gy and median dose to point A was 28 Gy (range: 16 ∼ 32 Gy) in 7 fractions. The median cumulative biologic effective dose (BED) at point A (EBRT + ICBT) was 88 Gy 10 (range:77 ∼ 94 Gy 10 ). The median cumulative BED at ICRU 38 reference point (EBRT + ICBT) was 131 Gy 3 (range: 122 ∼ 140 Gy 3 ) at point A, 109 Gy 3 (range:88∼ 125 Gy 3 ) at the rectum and 111 Gy 3 (range: 91 ∼ 123 Gy 3 ) at the urinary bladder. Cisplatin (60 mg/m 2 ) and 5-FU (1,000 mg/m 2 ) was administered intravenously at 2 weeks interval from the first day of radiation for median 5 (range:2 ∼ 6) cycles. The assessment was performed at 1 month after completion of radiation therapy by clinical examination and CT scan. The median follow-up time was 36 months (range:8∼ 50 months). The complete response rate after concurrent chemo radiation therapy was 93.3%. The 3-yr actuarial pelvic control rate was 87% and 3-yr actuarial overall survival and disease-free survival rate was 93% and 87%, respectively. The local failure rate was 13% and distant metastatic rate was 3.3%. The crude rate of minor hematologic

  6. Urethral stricture following high dose rate brachytherapy for prostate cancer

    International Nuclear Information System (INIS)

    Sullivan, Lisa; Williams, Scott G.; Tai, Keen Hun; Foroudi, Farshad; Cleeve, L.; Duchesne, Gillian M.

    2009-01-01

    Purpose: To evaluate the incidence, timing, nature and outcome of urethral strictures following high dose rate brachytherapy (HDRB) for prostate carcinoma. Methods and materials: Data from 474 patients with clinically localised prostate cancer treated with HDRB were analysed. Ninety percent received HDRB as a boost to external beam radiotherapy (HDRBB) and the remainder as monotherapy (HDRBM). Urethral strictures were graded according to the Common Terminology Criteria for Adverse Events v3.0. Results: At a median follow-up of 41 months, 38 patients (8%) were diagnosed with a urethral stricture (6-year actuarial risk 12%). Stricture location was bulbo-membranous (BM) urethra in 92.1%. The overall actuarial rate of grade 2 or more BM urethral stricture was estimated at 10.8% (95% CI 7.0-14.9%), with a median time to diagnosis of 22 months (range 10-68 months). All strictures were initially managed with either dilatation (n = 15) or optical urethrotomy (n = 20). Second line therapy was required in 17 cases (49%), third line in three cases (9%) and 1 patient open urethroplasty (grade 3 toxicity). Predictive factors on multivariate analysis were prior trans-urethral resection of prostate (hazard ratio (HR) 2.81, 95% CI 1.15-6.85, p = 0.023); hypertension (HR 2.83, 95% CI 1.37-5.85, p = 0.005); and dose per fraction used in HDR (HR for 1 Gy increase per fraction 1.33, 95% CI 1.08-1.64, p = 0.008). Conclusions: BM urethral strictures are the most common late grade 2 or more urinary toxicity following HDR brachytherapy for prostate cancer. Most are manageable with minimally invasive procedures. Both clinical and dosimetric factors appear to influence the risk of stricture formation.

  7. Implementation of High-Dose-Rate Brachytherapy and Androgen Deprivation in Patients With Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Lilleby, Wolfgang, E-mail: wolfgang.lilleby@ous-hf.no [Cancer Clinic, Oslo University Hospital, Norwegian Radiumhospital, Department of Radiotherapy and Oncology, Oslo (Norway); Tafjord, Gunnar; Raabe, Nils K. [Cancer Clinic, Oslo University Hospital, Norwegian Radiumhospital, Department of Radiotherapy and Oncology, Oslo (Norway)

    2012-07-01

    Purpose: To evaluate outcome (overall survival [OS], the actuarial 5-year cancer-specific survival [CSS], disease-free survival [DFS], biochemical failure-free survival [BFS]), complications and morbidity in patients treated with high-dose-rate brachytherapy (HDR-BT) boost and hormonal treatment with curative aims. Methods: Between 2004 and 2009, 275 prospectively followed pN0/N0M0 patients were included: 19 patients (7%) with T2, Gleason score 7 and prostate-specific antigen (PSA) <10 and 256 patients (93%) with T3 or Gleason score 8-10 or PSA >20 received multimodal treatment with conformal four-field radiotherapy (prostate/vesiculae 2 Gy Multiplication-Sign 25) combined with HDR-BT (iridium 192; prostate 10 Gy Multiplication-Sign 2) with long-term androgen deprivation therapy (ADT). Results: After a median observation time of 44.2 months (range, 10.4-90.5 months) 12 patients had relapsed clinically and/or biochemically and 10 patients were dead, of which 2 patients died from prostate cancer. Five-year estimates of BFS, CSS, DFS, and OS rates were 98.5%, 99.3%, 95.6%, and 96.3%, respectively. None of the patients with either Gleason score <8 or with intermediate risk profile had relapsed. The number of HDR-BT treatments was not related to outcome. Despite of age (median, 65.7 years; range, 45.7-77 years) and considerable pretreatment comorbidity in 39 of 275 patients, Genitourinary treatment-related morbidity was moderate with long-lasting Radiation Therapy Oncology Group Grade 2 voiding problems in 26 patients (9.5%) and occasionally mucous discharge in 20 patients (7%), none with Grade >2 for gastrointestinal at follow-up. Complications during implantations were related to pubic arch interference (4 patients) and lithotomy time, causing 2 patients to develop compartment syndrome. Conclusion: Despite still preliminary observations, our 5-year outcome estimates favor the implementation of high-dose-rate brachytherapy in high-risk patients combined with conformal

  8. Effective treatment of Stage I uterine papillary serous carcinoma with high dose-rate vaginal apex radiation (192Ir) and chemotherapy

    International Nuclear Information System (INIS)

    Turner, Bruce C.; Knisely, Jonathan P. S.; Kacinski, Barry M.; Haffty, Bruce G.; Gumbs, Andrew A.; Roberts, Kenneth B.; Frank, Alex H.; Peschel, Richard E.; Rutherford, Thomas J.; Edraki, Babak; Kohorn, Ernest I.; Chambers, Setsuko K.; Schwartz, Peter E.; Wilson, Lynn D.

    1998-01-01

    Purpose: Uterine papillary serous carcinoma (UPSC) is a morphologically distinct variant of endometrial carcinoma that is associated with a poor prognosis, high recurrence rate, frequent clinical understaging, and poor response to salvage treatment. We retrospectively analyzed local control, actuarial overall survival (OS), actuarial disease-free survival (DFS), salvage rate, and complications for patients with Federation International of Gynecology and Obstetrics (FIGO) (1988) Stage I UPSC. Methods and Materials: This retrospective analysis describes 38 patients with FIGO Stage I UPSC who were treated with the combinations of radiation therapy, chemotherapy, total abdominal hysterectomy, and bilateral salpingo-oophorectomy (TAH/BSO), with or without a surgical staging procedure. Twenty of 38 patients were treated with a combination of low dose-rate (LDR) uterine/vaginal brachytherapy using 226 Ra or 137 Cs and conventional whole-abdomen radiation therapy (WART) or whole-pelvic radiation therapy (WPRT). Of 20 patients (10%) in this treatment group, 2 received cisplatin chemotherapy. Eighteen patients were treated with high dose-rate (HDR) vaginal apex brachytherapy using 192 Ir with an afterloading device and cisplatin, doxorubicin, and cyclophosphamide (CAP) chemotherapy (5 of 18 patients). Only 6 of 20 UPSC patients treated with combination LDR uterine/vaginal brachytherapy and conventional external beam radiotherapy underwent complete surgical staging, consisting of TAH/BSO, pelvic/para-aortic lymph node sampling, omentectomy, and peritoneal fluid analysis, compared to 15 of 18 patients treated with HDR vaginal apex brachytherapy. Results: The 5-year actuarial OS for patients with complete surgical staging and adjuvant radiation/chemotherapy treatment was 100% vs. 61% for patients without complete staging (p = 0.002). The 5-year actuarial OS for all Stage I UPSC patients treated with postoperative HDR vaginal apex brachytherapy and systemic chemotherapy was 94

  9. Post-operative high dose rate vaginal apex brachytherapy in stage I endometrial adenocarcinoma

    International Nuclear Information System (INIS)

    Gumbs, A.A.; Turner, B.C.; Knisely, J.P.S.; Kacinski, B.M.; Roberts, K.B.; Peschel, R.E.; Haffty, B.G.; Rutherford, T.J.; Edraki, B.; Schwartz, P.E.; Wilson, L.D.

    1996-01-01

    Purpose/Objective: Patients with Stage I endometrial adenocarcinoma have traditionally been treated with total abdominal hysterectomy/bilateral salpingo-oophorectomy and radiation. The reported incidence of local recurrence in surgically treated patients with FIGO Stage IA or IB endometrial adenocarcinoma is 4-10% at 2 years. Combined surgery and radiation has resulted in a reduction of recurrence to 2-6%. We report the presentation, actuarial survival, actuarial rate of local failure, salvage rate, and complications for patients undergoing high dose rate (HDR) vaginal apex brachytherapy following surgery. Materials and Methods: Between 1985 and 1994 a total of 286 patients with FIGO Stage I endometrioid uterine adenocarcinoma were treated with HDR Ir-192 vaginal apex brachytherapy alone to a total dose of 21 Gy in 3 fractions at 0.5 cm from the vaginal mucosa. The pathologic stage by treatment group was IA: 31%, IB: 68%, and IC: 1%. The histologic grade of the patient's tumors was grade 1: 69%, 2: 29%, and 3: 2% of patients. The median time from surgery to radiation was 34 days (range 14-66 days). The median follow-up for 286 patients with Stage IA (92 patients), IB (190 patients), and IC (4 patients) was respectively, 37, 35 and 40 months (2 patients lost to follow-up prior to 6 months). Results: Patients presented with vaginal bleeding (94%) or abnormal pap smear (6%) at a median age for Stage IA and IB, of 54 and 63 years, respectively (range 32-88). The 5-year overall actuarial survival rate was 94.5%. The 5-year actuarial survival rate by histologic grade was 97.5% and 91.5% for FIGO grade 1 and 2, respectively (p=.011). The 5-year actuarial survival rate by depth of myometrial invasion was 99.0% and 92.5% for Stage IA and IB, respectively (p=.029). Median overall time to failure is 19.5 months (range 10-36 months). The 5-year actuarial rate of local failure was 4.5%. The overall failure rate in our study group was 2.8% (8 patients), local failure only 1

  10. Interstitial prostate brachytherapy. LDR-PDR-HDR

    International Nuclear Information System (INIS)

    Kovacs, Gyoergy; Hoskin, Peter

    2013-01-01

    The first comprehensive overview of interstitial brachytherapy for the management of local or locally advanced prostate cancer. Written by an interdisciplinary team who have been responsible for the successful GEC-ESTRO/EAU Teaching Course. Discusses in detail patient selection, the results of different methods, the role of imaging, and medical physics issues. Prostate brachytherapy has been the subject of heated debate among surgeons and the proponents of the various brachytherapy methods. This very first interdisciplinary book on the subject provides a comprehensive overview of innovations in low dose rate (LDR), high dose rate (HDR), and pulsed dose rate (PDR) interstitial brachytherapy for the management of local or locally advanced prostate cancer. In addition to detailed chapters on patient selection and the use of imaging in diagnostics, treatment guidance, and implantation control, background chapters are included on related medical physics issues such as treatment planning and quality assurance. The results obtained with the different treatment options and the difficult task of salvage treatment are fully discussed. All chapters have been written by internationally recognized experts in their fields who for more than a decade have formed the teaching staff responsible for the successful GEC-ESTRO/EAU Prostate Brachytherapy Teaching Course. This book will be invaluable in informing residents and others of the scientific background and potential of modern prostate brachytherapy. It will also prove a useful source of up-to-date information for those who specialize in prostate brachytherapy or intend to start an interstitial brachytherapy service.

  11. Coexistence Mechanism for Colocated HDR/LDR WPANs Air Interfaces

    Directory of Open Access Journals (Sweden)

    Prasad Ramjee

    2010-01-01

    Full Text Available This paper addresses the issues of interference management among Low Data Rate (LDR and High Data Rate (HDR WPAN air interfaces that are located in close-proximity (up to 10 cm and eventually on the same multimode device. After showing the noticeable performance degradation in terms of Bit Error Rate (BER and goodput due to the out-of-band interference of an HDR air interface over an LDR air interface, the paper presents a novel coexistence mechanism, named Alternating Wireless Activity (AWA, which is shown to greatly improve the performance in terms of goodput of the most interference vulnerable air interface (i.e., the LDR air interface. The main difference of the proposed mechanism with respect to other collaborative mechanisms based on time-scheduling is that it synchronizes the transmission of the LDR and HDR WPANs at the superframe level instead of packet level. Advantages and limitations of this choice are presented in the paper. Furthermore the functionalities of the AWA mechanism are positioned in a common protocol layer over the Medium Access Control (MAC sublayers of the HDR and LDR devices and it can be used with any standard whose MAC is based on a superframe structure.

  12. SU-E-J-93: Parametrisation of Dose to the Mucosa of the Anterior Rectal Wall in Transrectal Ultrasound Guided High-Dose-Rate Brachytherapy of the Prostate

    Energy Technology Data Exchange (ETDEWEB)

    Aitkenhead, A; Hamlett, L; Wood, D; Choudhury, A [The Christie Hospital NHS Foundation Trust, Manchester, Greater Manchester (United Kingdom)

    2014-06-01

    Purpose: In high-dose-rate (HDR) brachytherapy of the prostate, radiation is delivered from a number of radioactive sources which are inserted via catheter into the target volume. The rectal mucosa also receives dose during the treatment, which may lead to late toxicity effects. To allow possible links between rectal dose and toxicity to be investigated, suitable methods of parametrising the rectal dose are needed. Methods: During treatment of a series of 95 patients, anatomy and catheter locations were monitored by transrectal ultrasound, and target volume positions were contoured on the ultrasound scan by the therapist. The anterior rectal mucosal wall was identified by contouring the transrectal ultrasound balloon within the ultrasound scan. Source positions and dwell times, along with the dose delivered to the patient were computed using the Oncentra Prostate treatment planning system (TPS). Data for the series of patients were exported from the TPS in Dicom format, and a series of parametrisation methods were developed in a Matlab environment to assess the rectal dose. Results: Contours of the anterior rectal mucosa were voxelised within Matlab to allow the dose to the rectal mucosa to be analysed directly from the 3D dose grid. Dose parametrisations based on dose-surface (DSH) and dose-line (DLH) histograms were obtained. Both lateral and longitudinal extents of the mucosal dose were parametrised using dose-line histograms in the relevant directions. Conclusion: We have developed a series of dose parametrisations for quantifying the dose to the rectal mucosa during HDR prostate brachytherapy which are suitable for future studies investigating potential associations between mucosal dose and late toxicity effects. The geometry of the transrectal probe standardises the rectal anatomy, making this treatment technique particularly suited to studies of this nature.

  13. SU-E-J-93: Parametrisation of Dose to the Mucosa of the Anterior Rectal Wall in Transrectal Ultrasound Guided High-Dose-Rate Brachytherapy of the Prostate

    International Nuclear Information System (INIS)

    Aitkenhead, A; Hamlett, L; Wood, D; Choudhury, A

    2014-01-01

    Purpose: In high-dose-rate (HDR) brachytherapy of the prostate, radiation is delivered from a number of radioactive sources which are inserted via catheter into the target volume. The rectal mucosa also receives dose during the treatment, which may lead to late toxicity effects. To allow possible links between rectal dose and toxicity to be investigated, suitable methods of parametrising the rectal dose are needed. Methods: During treatment of a series of 95 patients, anatomy and catheter locations were monitored by transrectal ultrasound, and target volume positions were contoured on the ultrasound scan by the therapist. The anterior rectal mucosal wall was identified by contouring the transrectal ultrasound balloon within the ultrasound scan. Source positions and dwell times, along with the dose delivered to the patient were computed using the Oncentra Prostate treatment planning system (TPS). Data for the series of patients were exported from the TPS in Dicom format, and a series of parametrisation methods were developed in a Matlab environment to assess the rectal dose. Results: Contours of the anterior rectal mucosa were voxelised within Matlab to allow the dose to the rectal mucosa to be analysed directly from the 3D dose grid. Dose parametrisations based on dose-surface (DSH) and dose-line (DLH) histograms were obtained. Both lateral and longitudinal extents of the mucosal dose were parametrised using dose-line histograms in the relevant directions. Conclusion: We have developed a series of dose parametrisations for quantifying the dose to the rectal mucosa during HDR prostate brachytherapy which are suitable for future studies investigating potential associations between mucosal dose and late toxicity effects. The geometry of the transrectal probe standardises the rectal anatomy, making this treatment technique particularly suited to studies of this nature

  14. Treatment of skin carcinomas of the face by high-dose-rate brachytherapy and custom-made surface molds

    International Nuclear Information System (INIS)

    Guix, Benjamin; Finestres, Fernando; Tello, Jose-Ignacio; Palma, Cesar; Martinez, Antonio; Guix, Jose-Ramon; Guix, Ricardo

    2000-01-01

    Purpose: To analyze the results obtained in a prospective group of patients with basal or squamous cell skin carcinomas of the face treated by high-dose-rate (HDR) brachytherapy via custom-made surface molds. Methods and Materials: A total of 136 patients with basal or squamous cell carcinomas of the face were treated between March 1992 and March 1997 by surface molds and HDR brachytherapy with iridium-192. Nineteen patients were treated with standard Brock applicators and 117 patients with custom-made polymethyl methacrylate applicators, built over a plaster mold obtained of the patient's face. Minimum dose administered to the tumor was 6000 to 6500 cGy in 33 to 36 fractions at 180 cGy/fraction in lesions of up to 4 cm. Lesions greater than 4 cm were boosted up to 7500-8000 cGy after a 3-week pause. Results: With the custom-made surface molds, the dose distribution was uniform in the surface of the skin and at 5 mm depth in the whole area of the applicator. Differences between the areas of maximum and minimum dose at this depth never reached values higher than 5% of the prescribed dose. At the edges of the custom-made molds dose gradient was sharp, with the detected dose at 5 mm from the applicator being negligible. All the patients were complete responders. There were 3 local recurrences, 1/73 patients treated for primary tumor and 2/63 patients treated for recurrent tumor. Actuarial local control at 5 years for all patients was 98%, for those patients with primary tumors 99%, and for recurrent patients 87%. The treatment tolerance was excellent in all cases. No severe, early, or late, complications were detected. Conclusions: Radiotherapy is a highly effective treatment of skin carcinomas of the face. Custom-made molds, to be used in conjunction with HDR brachytherapy equipment, make possible a uniform dose distribution, with a sharp dose gradient in the limits of applicators. Custom-made surface molds are easy and safe to use, and they fit very accurately for

  15. Automated high-dose rate brachytherapy treatment planning for a single-channel vaginal cylinder applicator

    Science.gov (United States)

    Zhou, Yuhong; Klages, Peter; Tan, Jun; Chi, Yujie; Stojadinovic, Strahinja; Yang, Ming; Hrycushko, Brian; Medin, Paul; Pompos, Arnold; Jiang, Steve; Albuquerque, Kevin; Jia, Xun

    2017-06-01

    High dose rate (HDR) brachytherapy treatment planning is conventionally performed manually and/or with aids of preplanned templates. In general, the standard of care would be elevated by conducting an automated process to improve treatment planning efficiency, eliminate human error, and reduce plan quality variations. Thus, our group is developing AutoBrachy, an automated HDR brachytherapy planning suite of modules used to augment a clinical treatment planning system. This paper describes our proof-of-concept module for vaginal cylinder HDR planning that has been fully developed. After a patient CT scan is acquired, the cylinder applicator is automatically segmented using image-processing techniques. The target CTV is generated based on physician-specified treatment depth and length. Locations of the dose calculation point, apex point and vaginal surface point, as well as the central applicator channel coordinates, and the corresponding dwell positions are determined according to their geometric relationship with the applicator and written to a structure file. Dwell times are computed through iterative quadratic optimization techniques. The planning information is then transferred to the treatment planning system through a DICOM-RT interface. The entire process was tested for nine patients. The AutoBrachy cylindrical applicator module was able to generate treatment plans for these cases with clinical grade quality. Computation times varied between 1 and 3 min on an Intel Xeon CPU E3-1226 v3 processor. All geometric components in the automated treatment plans were generated accurately. The applicator channel tip positions agreed with the manually identified positions with submillimeter deviations and the channel orientations between the plans agreed within less than 1 degree. The automatically generated plans obtained clinically acceptable quality.

  16. Predictors of Toxicity After Image-guided High-dose-rate Interstitial Brachytherapy for Gynecologic Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Larissa J. [Department of Radiation Oncology, Brigham and Women' s Hospital and Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts (United States); Viswanathan, Akila N., E-mail: aviswanathan@lroc.harvard.edu [Department of Radiation Oncology, Brigham and Women' s Hospital and Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts (United States)

    2012-12-01

    Purpose: To identify predictors of grade 3-4 complications and grade 2-4 rectal toxicity after three-dimensional image-guided high-dose-rate (HDR) interstitial brachytherapy for gynecologic cancer. Methods and Materials: Records were reviewed for 51 women (22 with primary disease and 29 with recurrence) treated with HDR interstitial brachytherapy. A single interstitial insertion was performed with image guidance by computed tomography (n = 43) or magnetic resonance imaging (n = 8). The median delivered dose in equivalent 2-Gy fractions was 72.0 Gy (45 Gy for external-beam radiation therapy and 24 Gy for brachytherapy). Toxicity was reported according to the Common Toxicity Criteria for Adverse Events. Actuarial toxicity estimates were calculated by the Kaplan-Meier method. Results: At diagnosis, the median patient age was 62 years and the median tumor size was 3.8 cm. The median D90 and V100 were 71.4 Gy and 89.5%; the median D2cc for the bladder, rectum, and sigmoid were 64.6 Gy, 61.0 Gy, and 52.7 Gy, respectively. The actuarial rates of all grade 3-4 complications at 2 years were 20% gastrointestinal, 9% vaginal, 6% skin, 3% musculoskeletal, and 2% lymphatic. There were no grade 3-4 genitourinary complications and no grade 5 toxicities. Grade 2-4 rectal toxicity was observed in 10 patients, and grade 3-4 complications in 4; all cases were proctitis with the exception of 1 rectal fistula. D2cc for rectum was higher for patients with grade 2-4 (68 Gy vs 57 Gy for grade 0-1, P=.03) and grade 3-4 (73 Gy vs 58 Gy for grade 0-2, P=.02) rectal toxicity. The estimated dose that resulted in a 10% risk of grade 2-4 rectal toxicity was 61.8 Gy (95% confidence interval, 51.5-72.2 Gy). Discussion: Image-guided HDR interstitial brachytherapy results in acceptable toxicity for women with primary or recurrent gynecologic cancer. D2cc for the rectum is a reliable predictor of late rectal complications. Three-dimensional-based treatment planning should be performed to ensure

  17. Computed tomography-based treatment planning for high-dose-rate brachytherapy using the tandem and ring applicator: influence of applicator choice on organ dose and inter-fraction adaptive planning

    Directory of Open Access Journals (Sweden)

    Vishruta A. Dumane

    2017-06-01

    Full Text Available Three dimensional planning for high-dose-rate (HDR brachytherapy in cervical cancer has been highly recommended by consensus guidelines such as the American Brachytherapy Society (ABS and the Groupe Européen de Curiethérapie – European Society for Radiotherapy and Oncology (GEC-ESTRO. In this document, we describe our experience with computed tomography (CT-based planning using the tandem/ring applicator. We discuss the influence of applicator geometry on doses to organs at risk (OARs, namely the bladder, rectum, and sigmoid. Through example cases with dose prescribed to point A, we demonstrate how adaptive planning can help achieve constraints to the OARs as per guidelines.

  18. SU-E-T-375: Evaluation of a MapCHECK2(tm) Planar 2-D Diode Array for High-Dose-Rate Brachytherapy Treatment Delivery Verifications

    Energy Technology Data Exchange (ETDEWEB)

    Macey, N; Siebert, M; Shvydka, D; Parsai, E [University of Toledo Medical Center, Toledo, OH (United States)

    2015-06-15

    Purpose: Despite improvements of HDR brachytherapy delivery systems, verification of source position is still typically based on the length of the wire reeled out relative to the parked position. Yet, the majority of errors leading to medical events in HDR treatments continue to be classified as missed targets or wrong treatment sites. We investigate the feasibility of using dose maps acquired with a two-dimensional diode array to independently verify the source locations, dwell times, and dose during an HDR treatment. Methods: Custom correction factors were integrated into frame-by-frame raw counts recorded for a Varian VariSource™ HDR afterloader Ir-192 source located at various distances in air and in solid water from a MapCHECK2™ diode array. The resultant corrected counts were analyzed to determine the dwell position locations and doses delivered. The local maxima of polynomial equations fitted to the extracted dwell dose profiles provided the X and Y coordinates while the distance to the source was determined from evaluation of the full width at half maximum (FWHM). To verify the approach, the experiment was repeated as the source was moved through dwell positions at various distances along an inclined plane, mimicking a vaginal cylinder treatment. Results: Dose map analysis was utilized to provide the coordinates of the source and dose delivered over each dwell position. The accuracy in determining source dwell positions was found to be +/−1.0 mm of the preset values, and doses within +/−3% of those calculated by the BrachyVision™ treatment planning system for all measured distances. Conclusion: Frame-by-frame data furnished by a 2 -D diode array can be used to verify the dwell positions and doses delivered by the HDR source over the course of treatment. Our studies have verified that measurements provided by the MapCHECK2™ can be used as a routine QA tool for HDR treatment delivery verification.

  19. Quality control of the breast cancer treatments on Hdr brachytherapy with TLD-100

    Energy Technology Data Exchange (ETDEWEB)

    Torres H, F. [Universidad de Cordoba, Materials and Applied Physics Group, 230002 Monteria, Cordoba (Colombia); De la Espriella V, N. [Universidad de Cordoba, Grupo Avanzado de Materiales y Sistemas Complejos, 230002 Monteria, Cordoba (Colombia); Sanchez C, A., E-mail: franciscotorreshoyos@yahoo.com [Universidad de Cordoba, Departamento de Enfermeria, 230002 Monteria, Cordoba (Colombia)

    2014-07-01

    An anthropomorphic Phantom, a female trunk, was built with a natural bone structure and experimental material coated, glycerin and water-based material called JJT to build soft tissue equivalent to the muscle of human tissue, and a polymer (styrofoam) to build the lung as critical organ to simulate the treatment of breast cancer, with high dose rate brachytherapy (Hdr) and sources of Ir-192. The treatments were planned and calculated for the critical organ: Lung, and injury of 2 cm in diameter in breast with Micro Selectron Hdr system and the software Plato Brachytherapy V 14.1 of the Nucletron (Netherlands) which uses the standard protocol of radiotherapy for brachytherapy treatments. The dose experimentally measured with dosimeters TLD-100 LiF: Mg; Ti, which were previously calibrated, were placed in the same positions and bodies mentioned above, with less than 5% uncertainty. The reading dosimeters was carried out in a Harshaw TLD 4500. The results obtained for calculated treatments, using the standard simulator, and the experimental with TLD-100, show a high concordance, as they are on average a ± 1.1% making process becomes in a quality control of this type of treatments. (Author)

  20. Pre- and post-calculations for crack opening and leak rate experiments on piping components within the HDR-program

    International Nuclear Information System (INIS)

    Grebner, H.; Hoefler, A.; Hunger, H.

    1991-01-01

    In this paper calculations to experiments on leak opening and leak rates of piping components are presented. The experiments are performed at the HDR-facility at Karlstein/Germany and up to now straight pipes and pipe branches were considered. Numerical and experimental results are compared. (author)

  1. Dosimetric Evaluation of High-Dose-Rate Interstitial Brachytherapy Boost Treatments for Localized Prostate Cancer

    International Nuclear Information System (INIS)

    Froehlich, Georgina; Agoston, Peter; Loevey, Jozsef; Somogyi, Andras; Fodor, Janos; Polgar, Csaba; Major, Tibor

    2010-01-01

    Purpose: to quantitatively evaluate the dose distributions of high-dose-rate (HDR) prostate implants regarding target coverage, dose homogeneity, and dose to organs at risk. Material and methods: treatment plans of 174 implants were evaluated using cumulative dose-volume histograms (DVHs). The planning was based on transrectal ultrasound (US) imaging, and the prescribed dose (100%) was 10 Gy. The tolerance doses to rectum and urethra were 80% and 120%, respectively. Dose-volume parameters for target (V90, V100, V150, V200, D90, D min ) and quality indices (DNR [dose nonuniformity ratio], DHI [dose homogeneity index], CI [coverage index], COIN [conformal index]) were calculated. Maximum dose in reference points of rectum (D r ) and urethra (D u ), dose to volume of 2 cm 3 of the rectum (D 2ccm ), and 0.1 cm 3 and 1% of the urethra (D 0.1ccm and D1) were determined. Nonparametric correlation analysis was performed between these parameters. Results: the median number of needles was 16, the mean prostate volume (V p ) was 27.1 cm 3 . The mean V90, V100, V150, and V200 were 90%, 97%, 39% and 13%, respectively. The mean D90 was 109%, and the D min was 87%. The mean doses in rectum and urethra reference points were 75% and 119%, respectively. The mean volumetric doses were D 2ccm = 49% for the rectum, D 0.1ccm = 126%, and D1 = 140% for the urethra. The mean DNR was 0.37, while the DHI was 0.60. The mean COIN was 0.66. The Spearman rank order correlation coefficients for volume doses to rectum and urethra were R(D r , D 2ccm ) = 0.69, R(D u , D 0.1ccm ) = 0.64, R(D u , D1) = 0.23. Conclusion: US-based treatment plans for HDR prostate implants based on the real positions of catheters provided acceptable dose distributions. In the majority of the cases, the doses to urethra and rectum were kept below the defined tolerance levels. For rectum, the dose in reference points correlated well with dose-volume parameters. For urethra dose characterization, the use of D1 volumetric

  2. Dosimetric evaluation of high-dose-rate interstitial brachytherapy boost treatments for localized prostate cancer.

    Science.gov (United States)

    Fröhlich, Georgina; Agoston, Péter; Lövey, József; Somogyi, András; Fodor, János; Polgár, Csaba; Major, Tibor

    2010-07-01

    To quantitatively evaluate the dose distributions of high-dose-rate (HDR) prostate implants regarding target coverage, dose homogeneity, and dose to organs at risk. Treatment plans of 174 implants were evaluated using cumulative dose-volume histograms (DVHs). The planning was based on transrectal ultrasound (US) imaging, and the prescribed dose (100%) was 10 Gy. The tolerance doses to rectum and urethra were 80% and 120%, respectively. Dose-volume parameters for target (V90, V100, V150, V200, D90, D(min)) and quality indices (DNR [dose nonuniformity ratio], DHI [dose homogeneity index], CI [coverage index], COIN [conformal index]) were calculated. Maximum dose in reference points of rectum (D(r)) and urethra (D(u)), dose to volume of 2 cm(3) of the rectum (D(2ccm)), and 0.1 cm(3) and 1% of the urethra (D(0.1ccm) and D1) were determined. Nonparametric correlation analysis was performed between these parameters. The median number of needles was 16, the mean prostate volume (V(p)) was 27.1 cm(3). The mean V90, V100, V150, and V200 were 99%, 97%, 39%, and 13%, respectively. The mean D90 was 109%, and the D(min) was 87%. The mean doses in rectum and urethra reference points were 75% and 119%, respectively. The mean volumetric doses were D(2ccm) = 49% for the rectum, D(0.1ccm) = 126%, and D1 = 140% for the urethra. The mean DNR was 0.37, while the DHI was 0.60. The mean COIN was 0.66. The Spearman rank order correlation coefficients for volume doses to rectum and urethra were R(D(r),D(2ccm)) = 0.69, R(D(u),D0.(1ccm)) = 0.64, R(D(u),D1) = 0.23. US-based treatment plans for HDR prostate implants based on the real positions of catheters provided acceptable dose distributions. In the majority of the cases, the doses to urethra and rectum were kept below the defined tolerance levels. For rectum, the dose in reference points correlated well with dose-volume parameters. For urethra dose characterization, the use of D1 volumetric parameter is recommended.

  3. A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Qi Zhenyu; Deng Xiaowu; Cao Xinping; Huang Shaomin; Lerch, Michael; Rosenfeld, Anatoly [State Key Laboratory of Oncology in South China, Sun Yat-sen University Cancer Center, Guangzhou 510060 (China) and Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522 (Australia); State Key Laboratory of Oncology in South China, Sun Yat-sen University Cancer Center, Guangzhou 510060 (China); Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522 (Australia)

    2012-11-15

    Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in {sup 192}Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 {+-} 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1%{+-} 3.8% and -0.1%{+-} 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors.

  4. Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera

    International Nuclear Information System (INIS)

    Duan, Jun; Macey, Daniel J.; Pareek, Prem N.; Brezovich, Ivan A.

    2001-01-01

    We investigated a pinhole imaging system for independent in vivo monitoring and verification of high dose rate (HDR) brachytherapy treatment. The system consists of a high-resolution pinhole collimator, an x-ray fluoroscope, and a standard radiographic screen-film combination. Autofluoroscopy provides real-time images of the in vivo Ir-192 HDR source for monitoring the source location and movement, whereas autoradiography generates a permanent record of source positions on film. Dual-pinhole autoradiographs render stereo-shifted source images that can be used to reconstruct the source dwell positions in three dimensions. The dynamic range and spatial resolution of the system were studied with a polystyrene phantom using a range of source strengths and dwell times. For the range of source activity used in HDR brachytherapy, a 0.5 mm diameter pinhole produced sharp fluoroscopic images of the source within the dynamic range of the fluoroscope. With a source-to-film distance of 35 cm and a 400 speed screen-film combination, the same pinhole yielded well recognizable images of a 281.2 GBq (7.60 Ci) Ir-192 source for dwell times in the typical clinical range of 2 to 400 s. This 0.5 mm diameter pinhole could clearly resolve source positions separated by lateral displacements as small as 1 mm. Using a simple reconstruction algorithm, dwell positions in a phantom were derived from stereo-shifted dual-pinhole images and compared to the known positions. The agreement was better than 1 mm. A preliminary study of a patient undergoing HDR treatment for cervical cancer suggests that the imaging method is clinically feasible. Based on these studies we believe that the pinhole imaging method is capable of providing independent and reliable real-time monitoring and verification for HDR brachytherapy

  5. An analysis of acute complications and perioperative morbidity from high dose rate brachytherapy in the treatment of gynecological malignancies

    International Nuclear Information System (INIS)

    Sarkaria, Jann N.; Petereit, Daniel G.; Kinsella, Timothy J.; Buchler, Dolores A.

    1995-01-01

    Purpose: To evaluate the acute morbidity and mortality for high dose-rate (HDR) brachytherapy performed in an outpatient setting in the treatment of gynecological malignancies, and to identify possible risk factors for adverse outcomes. Materials and Methods: One hundred seventy-one patients with cervical (n=129) or uterine (n=42) carcinoma with an intact uterus were evaluated and treated from August 1989 through December 1994, with at least part of their therapy delivered with intracavitary HDR 192 Ir radiation. A total of 830 ICR insertions were performed with greater than 95% done on an outpatient basis under heavy intravenous sedation using fentanyl and midazolam. Perioperative morbidity and mortality were recorded for any event occurring within 30 days of the completion of therapy. Anesthesia risk was evaluated retrospectively in all patients based on the American Society of Anesthesiologists' (ASA) Physical Class System. Results: The uterine patients, many treated with radiation alone because of morbid obesity or medical inoperability, had a significantly higher perioperative morbidity and mortality rate as compared to the cervix patient cohort. Thirteen of the 42 (31%) uterine patients and 8 of the 129 (6%) cervix patients required hospitalization within 30 days of treatment completion (p 2) experienced greater morbidity and mortality, while the best predictor of complications in the cervix patients was age greater than 70 years. For the entire cohort of patients, no correlation was found between the 30 day morbidity and mortality and the doses of fentanyl and midazolam used or the length of the procedure. Conclusions: The acute complication rate from HDR brachytherapy performed on an outpatient basis with heavy intravenous sedation is acceptable for the great majority of patients who present for treatment. However, the high morbidity and mortality experienced by certain patient cohorts suggests that careful assessment of the risk/benefit ratio for treatment

  6. Applicability and dosimetric impact of ultrasound-based preplanning in high-dose-rate brachytherapy of prostate cancer

    International Nuclear Information System (INIS)

    Aebersold, D.M.; Isaak, B.; Behrensmeier, F.; Kolotas, C.; Mini, R.; Greiner, R.H.; Thalmann, G.; Kranzbuehler, H.

    2004-01-01

    Background and purpose: analyses of permanent brachytherapy seed implants of the prostate have demonstrated that the use of a preplan may lead to a considerable decrease of dosimetric implant quality. The authors aimed to determine whether the same drawbacks of preplanning also apply to high-dose-rate (HDR) brachytherapy. Patients and methods: 15 patients who underwent two separate HDR brachytherapy implants in addition to external-beam radiation therapy for advanced prostate cancer were analyzed. A pretherapeutic transrectal ultrasound was performed in all patients to generate a preplan for the first brachytherapy implant. For the second brachytherapy, a subset of patients were treated by preplans based on the ultrasound from the first brachytherapy implant. Preplans were compared with the respective postplans assessing the following parameters: coverage index, minimum target dose, homogeneity index, and dose exposure of organs at risk. The prostate geometries (volume, width, height, length) were compared as well. Results: at the first brachytherapy, the matching between the preplan and actual implant geometry was sufficient in 47% of the patients, and the preplan could be applied. The dosimetric implant quality decreased considerably: the mean coverage differed by -0.11, the mean minimum target dose by -0.15, the mean homogeneity index by -0.09. The exposure of organs at risk was not substantially altered. At the second brachytherapy, all patients could be treated by the preplan; the differences between the implant quality parameters were less pronounced. The changes of prostate geometry between preplans and postplans were considerable, the differences in volume ranging from -8.0 to 13.8 cm 3 and in dimensions (width, height, length) from -1.1 to 1.0 cm. Conclusion: preplanning in HDR brachytherapy of the prostate is associated with a substantial decrease of dosimetric implant quality, when the preplan is based on a pretherapeutic ultrasound. The implant quality

  7. Attenuation measurements show that the presence of a TachoSil surgical patch will not compromise target irradiation in intra-operative electron radiation therapy or high-dose-rate brachytherapy

    International Nuclear Information System (INIS)

    Sarmento, Sandra; Costa, Filipa; Pereira, Alexandre; Lencart, Joana; Dias, Anabela; Cunha, Luís; Sousa, Olga; Silva, José Pedro; Santos, Lúcio

    2015-01-01

    Surgery of locally advanced and/or recurrent rectal cancer can be complemented with intra-operative electron radiation therapy (IOERT) to deliver a single dose of radiation directly to the unresectable margins, while sparing nearby sensitive organs/structures. Haemorrhages may occur and can affect the dose distribution, leading to an incorrect target irradiation. The TachoSil (TS) surgical patch, when activated, creates a fibrin clot at the surgical site to achieve haemostasis. The aim of this work was to determine the effect of TS on the dose distribution, and ascertain whether it could be used in combination with IOERT. This characterization was extended to include high dose rate (HDR) intraoperative brachytherapy, which is sometimes used at other institutions instead of IOERT. CT images of the TS patch were acquired for initial characterization. Dosimetric measurements were performed in a water tank phantom, using a conventional LINAC with a hard-docking system of cylindrical applicators. Percentage Depth Dose (PDD) curves were obtained, and measurements made at the depth of dose maximum for the three clinically used electron energies (6, 9 and 12MeV), first without any attenuator and then with the activated patch of TS completely covering the tip of the IOERT applicator. For HDR brachytherapy, a measurement setup was improvised using a solid water phantom and a Farmer ionization chamber. Our measurements show that the attenuation of a TachoSil patch is negligible, both for high energy electron beams (6 to 12MeV), and for a HDR 192 Ir brachytherapy source. Our results cannot be extrapolated to lower beam energies such as 50 kVp X-rays, which are sometimes used for breast IORT. The TachoSil surgical patch can be used in IORT procedures using 6MeV electron energies or higher, or HDR 192 Ir brachytherapy

  8. Attenuation measurements show that the presence of a TachoSil surgical patch will not compromise target irradiation in intra-operative electron radiation therapy or high-dose-rate brachytherapy.

    Science.gov (United States)

    Sarmento, Sandra; Costa, Filipa; Pereira, Alexandre; Lencart, Joana; Dias, Anabela; Cunha, Luís; Sousa, Olga; Silva, José Pedro; Santos, Lúcio

    2015-01-09

    Surgery of locally advanced and/or recurrent rectal cancer can be complemented with intra-operative electron radiation therapy (IOERT) to deliver a single dose of radiation directly to the unresectable margins, while sparing nearby sensitive organs/structures. Haemorrhages may occur and can affect the dose distribution, leading to an incorrect target irradiation. The TachoSil (TS) surgical patch, when activated, creates a fibrin clot at the surgical site to achieve haemostasis. The aim of this work was to determine the effect of TS on the dose distribution, and ascertain whether it could be used in combination with IOERT. This characterization was extended to include high dose rate (HDR) intraoperative brachytherapy, which is sometimes used at other institutions instead of IOERT. CT images of the TS patch were acquired for initial characterization. Dosimetric measurements were performed in a water tank phantom, using a conventional LINAC with a hard-docking system of cylindrical applicators. Percentage Depth Dose (PDD) curves were obtained, and measurements made at the depth of dose maximum for the three clinically used electron energies (6, 9 and 12MeV), first without any attenuator and then with the activated patch of TS completely covering the tip of the IOERT applicator. For HDR brachytherapy, a measurement setup was improvised using a solid water phantom and a Farmer ionization chamber. Our measurements show that the attenuation of a TachoSil patch is negligible, both for high energy electron beams (6 to 12MeV), and for a HDR (192)Ir brachytherapy source. Our results cannot be extrapolated to lower beam energies such as 50 kVp X-rays, which are sometimes used for breast IORT. The TachoSil surgical patch can be used in IORT procedures using 6MeV electron energies or higher, or HDR (192)Ir brachytherapy.

  9. The Curie–Da Vinci Connection: 5-Years' Experience With Laparoscopic (Robot-Assisted) Implantation for High-Dose-Rate Brachytherapy of Solitary T2 Bladder Tumors

    Energy Technology Data Exchange (ETDEWEB)

    Steen-Banasik, Elzbieta M. van der, E-mail: E.vanderSteen-Banasik@radiotherapiegroep.nl [Radiotherapiegroep, Arnhem (Netherlands); Smits, Geert A.H.J. [Department of Urology, Rijnstate Hospital, Arnhem (Netherlands); Oosterveld, Bernard J.; Janssen, Theo; Visser, Andries G. [Radiotherapiegroep, Arnhem (Netherlands)

    2016-08-01

    Purpose: To report experience and early results of laparoscopic implantation for interstitial brachytherapy (BT) of solitary bladder tumors and the feasibility of a high-dose-rate (HDR) schedule. Methods and Materials: From December 2009 to April 2015, 57 patients with a T2 solitary bladder tumor were treated in Arnhem with transurethral bladder resection followed by external beam irradiation, applied to the bladder and regional iliac lymph nodes, 40 Gy in 20 fractions, 5 fractions per week, and within 1 week interstitial HDR BT, in selected cases combined with partial cystectomy and lymph node dissection. The BT catheters were placed via a transabdominal approach with robotic assistance from a Da Vinci robot after a successful initial experience with a nonrobotic laparoscopic approach. The fraction schedule for HDR was 10 fractions of 2.5 Gy, 3 fractions per day. This was calculated to be equivalent to a reference low-dose-rate schedule of 30 Gy in 60 hours. Data for oncologic outcomes and toxicity (Common Toxicity Criteria version 4) were prospectively collected. Results: These modifications resulted in an average postoperative hospitalization of 6 days, minimal blood loss, and no wound healing problems. Two patients had severe acute toxicity: 1 pulmonary embolism grade 4 and 1 cardiac death. Late toxicity was mild (n=2 urogenital grade 3 toxicity). The median follow-up was 2 years. Using cumulative incidence competing risk analysis, the 2-year overall, disease-free, and disease-specific survival and local control rates were 59%, 71%, 87%, and 82%, respectively. Conclusions: The benefits of minimally invasive surgery for implantation of BT catheters and the feasibility of HDR BT in bladder cancer are documented. The patient outcome and adverse events are comparable to the best results published for a bladder-sparing approach.

  10. HDR Brachytherapy in the Management of High-Risk Prostate Cancer

    Directory of Open Access Journals (Sweden)

    Susan Masson

    2012-01-01

    Full Text Available High-dose-rate (HDR brachytherapy is used with increasing frequency for the treatment of prostate cancer. It is a technique which allows delivery of large individual fractions to the prostate without exposing adjacent normal tissues to unacceptable toxicity. This approach is particularly favourable in prostate cancer where tumours are highly sensitive to dose escalation and to increases in radiotherapy fraction size, due to the unique radiobiological behaviour of prostate cancers in contrast with other malignancies. In this paper we discuss the rationale and the increasing body of clinical evidence for the use of this technique in patients with high-risk prostate cancer, where it is combined with external beam radiotherapy. We highlight practical aspects of delivering treatment and discuss toxicity and limitations, with particular reference to current practice in the United Kingdom.

  11. SU-F-T-63: Dosimetric Relevance of the Valencia and Leipzig HDR Applicators Plastic Cap

    Energy Technology Data Exchange (ETDEWEB)

    Granero, D [ERESA-Hospital General Universitario, Valencia (Spain); Candela-Juan, C [National Dosimetry Centre (CND), Valencia (Spain); Vijande, J; Ballester, F [University of Valencia, Burjassot (Spain); Perez-Calatayud, J [Hospital La Fe, Valencia (Spain); Jacob, D; Mourtada, F [Helen F. Graham Cancer Center, Christiana Care Health System, Newark, DE (United States)

    2016-06-15

    Purpose: Utilization of HDR brachytherapy treatment of skin lesions using collimated applicators, such as the Valencia or Leipzig is increasing. These applicators are made of cup-shaped tungsten material in order to focalize the radiation into the lesion and to protect nearby tissues. These applicators have an attachable plastic cap that removes secondary electrons generated in the applicator and flattens the treatment surface. The purpose of this study is to examine the dosimetric impact of this cap, and the effect if the cap is not placed during the HDR fraction delivery. Methods: Monte Carlo simulations have been done using the code Geant4 for the Valencia and Leipzig applicators. Dose rate distributions have been obtained for the applicators with and without the plastic cap. An experimental study using EBT3 radiochromic film has been realized in order to verify the Monte Carlo results. Results: The Monte Carlo simulations show that absorbed dose in the first millimeter of skin can increase up to 180% for the Valencia applicator if the plastic cap is absent and up to 1500% for the Leipzig applicators. At deeper distances the increase of dose is smaller being about 10–15%. Conclusion: Important differences have been found if the plastic cap of the applicators is absent in the treatment producing an overdosage in the skin. The user should have a checklist to remind him check always before HDR fraction delivery to insure the plastic cap is placed on the applicator. This work was supported in part by Generalitat Valenciana under Project PROMETEOII/2013/010, by the Spanish Government under Project No. FIS2013-42156, and by a research agreement with Elekta Brachytherapy, Veenendaal, The Netherlands.

  12. Fricke gel-layer dosimetry in HDR brachytherapy

    International Nuclear Information System (INIS)

    Gambarini, G.; Negri, A.; Carrara, M.; Marchesini, R.

    2008-01-01

    Full text: In the last decade, technological improvements in radiotherapy have been significant and consequently the use and importance of radiotherapy in cancer treatment have increased greatly. In brachytherapy, new possibilities have been opened by the impressive progresses in 3D imaging, by the development of sophisticated techniques for modern afterloaders and by the constantly increasing speed and capacity of computers. However, these methodological improvements require corresponding improvements in the dosimetry methods, in order to ensure that the values calculated with computer treatment planning systems, adopted in the clinical praxis, agree with the delivered dose distributions. Fricke gel-layer dosimeters (FGLD) are under study by our group as a reliable alternative to films, semiconductors arrays or thermoluminescent dosimeters (TLDs). In the last years, we have significantly improved this technique by defining the FGLD best chemical composition, by optimizing the image acquisition assessment and by developing a dedicated software for image analysis. In this study, experimental measurements of planar dose distributions of a clinical 192 Ir source (Microselectron HDR, Nucletron) obtained by irradiating a series of piled-up FGL dosimeters in a tissue-equivalent phantom are presented. The obtained results were in accordance to TLD measurements and to treatment planning system (Plato, Nucletron) calculations. FGLD have proven to be a reliable tool to achieve HDR brachytherapy dose distribution measurements

  13. Interactive, multi-modality image registrations for combined MRI/MRSI-planned HDR prostate brachytherapy

    Directory of Open Access Journals (Sweden)

    Galen Reed

    2011-03-01

    Full Text Available Purpose: This study presents the steps and criteria involved in the series of image registrations used clinically during the planning and dose delivery of focal high dose-rate (HDR brachytherapy of the prostate. Material and methods: Three imaging modalities – Magnetic Resonance Imaging (MRI, Magnetic Resonance Spectroscopic Imaging (MRSI, and Computed Tomography (CT – were used at different steps during the process. MRSI is used for identification of dominant intraprosatic lesions (DIL. A series of rigid and nonrigid transformations were applied to the data to correct for endorectal-coil-induced deformations and for alignment with the planning CT. Mutual information was calculated as a morphing metric. An inverse planning optimization algorithm was applied to boost dose to the DIL while providing protection to the urethra, penile bulb, rectum, and bladder. Six prostate cancer patients were treated using this protocol. Results: The morphing algorithm successfully modeled the probe-induced prostatic distortion. Mutual information calculated between the morphed images and images acquired without the endorectal probe showed a significant (p = 0.0071 increase to that calculated between the unmorphed images and images acquired without the endorectal probe. Both mutual information and visual inspection serve as effective diagnostics of image morphing. The entire procedure adds less than thirty minutes to the treatment planning. Conclusion: This work demonstrates the utility of image transformations and registrations to HDR brachytherapy of prostate cancer.

  14. High-Dose-Rate Brachytherapy of a Single Implant With Two Fractions Combined With External Beam Radiotherapy for Hormone-Naive Prostate Cancer

    International Nuclear Information System (INIS)

    Sato, Morio; Mori, Takashi; Shirai, Shintaro; Kishi, Kazushi; Inagaki, Takeshi; Hara, Isao

    2008-01-01

    Purpose: To evaluate the preliminary outcomes of high-dose-rate (HDR) brachytherapy of a single implant with two fractions and external beam radiotherapy (EBRT) for hormone-naive prostate cancer. Methods and Materials: Between March 2000 and Sept 2003, a total of 53 patients with tumor Stage T1c-T3b N0 M0 prostate cancer were treated with HDR brachytherapy boost doses (7.5 Gy/fraction) and 50-Gy EBRT during a 5.5-week period. Median follow-up was 61 months. Patients were divided into groups with localized (T1c-T2b) and advanced disease (T3a-T3b). We used the American Society for Therapeutic Radiology and Oncology (ASTRO) definition for biochemical failure. According to recommendations of the Radiation Therapy Oncology Group-ASTRO Phoenix Consensus Conference, biochemical failure-free control rates (BF-FCRs) at 3 years were investigated as 2 years short of the median follow-up. Results: Between April 2000 and Sept 2007, Common Terminology Criteria for Adverse Events Version 2.0 late Grade 2 genitourinary and gastrointestinal toxicity rates were 0% and 3.8%, respectively. Erectile preservation was 25% at 5 years. Overall survival was 88.1% and cause-specific survival was 100%. At 3 years, ASTRO BF-FCRs of the localized and advanced groups were 100% and 42%, respectively (p = 0.001). Conclusions: The HDR brachytherapy of a single implant with two fractions plus EBRT is effective in treating patients with localized hormone-naive prostate cancer, with the least genitourinary and gastrointestinal toxicities; however, longer median BF-FCR follow-up is required to assess these findings

  15. In vivo assessment of the gastric mucosal tolerance dose after single fraction, small volume irradiation of liver malignancies by computed tomography-guided, high-dose-rate brachytherapy

    International Nuclear Information System (INIS)

    Streitparth, Florian; Pech, Maciej; Boehmig, Michael; Ruehl, Ricarda; Peters, Nils; Wieners, Gero; Steinberg, Johannes; Lopez-Haenninen, Enrique; Felix, Roland; Wust, Peter; Ricke, Jens

    2006-01-01

    Purpose: The aim of this study was to assess the tolerance dose of gastric mucosa for single-fraction computed tomography (CT)-guided, high-dose-rate (HDR) brachytherapy of liver malignancies. Methods and Materials: A total of 33 patients treated by CT-guided HDR brachytherapy of liver malignancies in segments II and/or III were included. Dose planning was performed upon a three-dimensional CT data set acquired after percutaneous applicator positioning. All patients received gastric protection post-treatment. For further analysis, the contours of the gastric wall were defined in every CT slice using Brachyvision Software. Dose-volume histograms were calculated for each treatment and correlated with clinical data derived from questionnaires assessing Common Toxicity Criteria (CTC). All patients presenting symptoms of upper GI toxicity were examined endoscopically. Results: Summarizing all patients the minimum dose applied to 1 ml of the gastric wall (D 1ml ) ranged from 6.3 to 34.2 Gy; median, 14.3 Gy. Toxicity was present in 18 patients (55%). We found nausea in 16 (69%), emesis in 9 (27%), cramping in 13 (39%), weight loss in 12 (36%), gastritis in 4 (12%), and ulceration in 5 patients (15%). We found a threshold dose D 1ml of 11 Gy for general gastric toxicity and 15.5 Gy for gastric ulceration verified by an univariate analysis (p = 0.01). Conclusions: For a single fraction, small volume irradiation we found in the upper abdomen a threshold dose D 1ml of 15.5 Gy for the clinical endpoint ulceration of the gastric mucosa. This in vivo assessment is in accordance with previously published tolerance data

  16. Feasibility and early results of interstitial intensity-modulated HDR/PDR brachytherapy (IMBT) with/without complementary external-beam radiotherapy and extended surgery in recurrent pelvic colorectal cancer

    International Nuclear Information System (INIS)

    Tepel, J.; Bokelmann, F.; Faendrich, F.; Kremer, B.; Schmid, A.; Kovacs, G.; University Hospital of Schleswig-Holstein, Kiel

    2005-01-01

    Background: A new multimodality treatment concept consisting of extended resection and postoperative fractionated intensity-modulated interstitial brachytherapy (IMBT) was introduced for pelvic recurrence of colorectal carcinoma. Patients and Methods: 46 patients received extended resection and single plastic tubes were sutured directly onto the tumor bed. IMBT was started within 2 weeks postoperatively with a median dose of 24.5 Gy (5-35 Gy). Patients were treated either with high-dose-rate brachytherapy (HDR; n=23) or with pulsed-dose-rate brachytherapy (PDR; n=23). 25 patients received complementary 45-Gy external-beam irradiation (EBRT) to the pelvic region after explanting the plastic tubes. Results: Median follow-up was 20.6 months (7-107 months) and mean patient survival 25.7±25.8 months (median 17, range 1-107 months). After 5 years overall survival, disease-free survival and local control rate were 23%, 20% and 33%, significantly influenced by the resectional state. There was a trend in favor of PDR compared to HDR, which reached statistical significance in patients who had not received additional EBRT. Conclusion: The combination of extended surgery and postoperative interstitial IMBT is feasible and offers effective interdisciplinary treatment of recurrent colorectal cancer. In this small and inhomogeneous cohort of patients PDR seems to be more effective than HDR, particularly when application of complementary EBRT is not possible. None of the patients who required resection of distant metastasis survived >2 years in this study. (orig.)

  17. A multicentre audit of HDR/PDR brachytherapy absolute dosimetry in association with the INTERLACE trial (NCT015662405)

    Science.gov (United States)

    Díez, P.; Aird, E. G. A.; Sander, T.; Gouldstone, C. A.; Sharpe, P. H. G.; Lee, C. D.; Lowe, G.; Thomas, R. A. S.; Simnor, T.; Bownes, P.; Bidmead, M.; Gandon, L.; Eaton, D.; Palmer, A. L.

    2017-12-01

    A UK multicentre audit to evaluate HDR and PDR brachytherapy has been performed using alanine absolute dosimetry. This is the first national UK audit performing an absolute dose measurement at a clinically relevant distance (20 mm) from the source. It was performed in both INTERLACE (a phase III multicentre trial in cervical cancer) and non-INTERLACE brachytherapy centres treating gynaecological tumours. Forty-seven UK centres (including the National Physical Laboratory) were visited. A simulated line source was generated within each centre’s treatment planning system and dwell times calculated to deliver 10 Gy at 20 mm from the midpoint of the central dwell (representative of Point A of the Manchester system). The line source was delivered in a water-equivalent plastic phantom (Barts Solid Water) encased in blocks of PMMA (polymethyl methacrylate) and charge measured with an ion chamber at 3 positions (120° apart, 20 mm from the source). Absorbed dose was then measured with alanine at the same positions and averaged to reduce source positional uncertainties. Charge was also measured at 50 mm from the source (representative of Point B of the Manchester system). Source types included 46 HDR and PDR 192Ir sources, (7 Flexisource, 24 mHDR-v2, 12 GammaMed HDR Plus, 2 GammaMed PDR Plus, 1 VS2000) and 1 HDR 60Co source, (Co0.A86). Alanine measurements when compared to the centres’ calculated dose showed a mean difference (±SD) of  +1.1% (±1.4%) at 20 mm. Differences were also observed between source types and dose calculation algorithm. Ion chamber measurements demonstrated significant discrepancies between the three holes mainly due to positional variation of the source within the catheter (0.4%-4.9% maximum difference between two holes). This comprehensive audit of absolute dose to water from a simulated line source showed all centres could deliver the prescribed dose to within 5% maximum difference between measurement and calculation.

  18. SU-F-19A-10: Recalculation and Reporting Clinical HDR 192-Ir Head and Neck Dose Distributions Using Model Based Dose Calculation

    Energy Technology Data Exchange (ETDEWEB)

    Carlsson Tedgren, A [Linkoping University, Linkoping, Linkoping (Sweden); Persson, M; Nilsson, J [Karolinska hospital, Stockholm, Stockholm (Sweden)

    2014-06-15

    Purpose: To retrospectively re-calculate dose distributions for selected head and neck cancer patients, earlier treated with HDR 192Ir brachytherapy, using Monte Carlo (MC) simulations and compare results to distributions from the planning system derived using TG43 formalism. To study differences between dose to medium (as obtained with the MC code) and dose to water in medium as obtained through (1) ratios of stopping powers and (2) ratios of mass energy absorption coefficients between water and medium. Methods: The MC code Algebra was used to calculate dose distributions according to earlier actual treatment plans using anonymized plan data and CT images in DICOM format. Ratios of stopping power and mass energy absorption coefficients for water with various media obtained from 192-Ir spectra were used in toggling between dose to water and dose to media. Results: Differences between initial planned TG43 dose distributions and the doses to media calculated by MC are insignificant in the target volume. Differences are moderate (within 4–5 % at distances of 3–4 cm) but increase with distance and are most notable in bone and at the patient surface. Differences between dose to water and dose to medium are within 1-2% when using mass energy absorption coefficients to toggle between the two quantities but increase to above 10% for bone using stopping power ratios. Conclusion: MC predicts target doses for head and neck cancer patients in close agreement with TG43. MC yields improved dose estimations outside the target where a larger fraction of dose is from scattered photons. It is important with awareness and a clear reporting of absorbed dose values in using model based algorithms. Differences in bone media can exceed 10% depending on how dose to water in medium is defined.

  19. SU-F-19A-10: Recalculation and Reporting Clinical HDR 192-Ir Head and Neck Dose Distributions Using Model Based Dose Calculation

    International Nuclear Information System (INIS)

    Carlsson Tedgren, A; Persson, M; Nilsson, J

    2014-01-01

    Purpose: To retrospectively re-calculate dose distributions for selected head and neck cancer patients, earlier treated with HDR 192Ir brachytherapy, using Monte Carlo (MC) simulations and compare results to distributions from the planning system derived using TG43 formalism. To study differences between dose to medium (as obtained with the MC code) and dose to water in medium as obtained through (1) ratios of stopping powers and (2) ratios of mass energy absorption coefficients between water and medium. Methods: The MC code Algebra was used to calculate dose distributions according to earlier actual treatment plans using anonymized plan data and CT images in DICOM format. Ratios of stopping power and mass energy absorption coefficients for water with various media obtained from 192-Ir spectra were used in toggling between dose to water and dose to media. Results: Differences between initial planned TG43 dose distributions and the doses to media calculated by MC are insignificant in the target volume. Differences are moderate (within 4–5 % at distances of 3–4 cm) but increase with distance and are most notable in bone and at the patient surface. Differences between dose to water and dose to medium are within 1-2% when using mass energy absorption coefficients to toggle between the two quantities but increase to above 10% for bone using stopping power ratios. Conclusion: MC predicts target doses for head and neck cancer patients in close agreement with TG43. MC yields improved dose estimations outside the target where a larger fraction of dose is from scattered photons. It is important with awareness and a clear reporting of absorbed dose values in using model based algorithms. Differences in bone media can exceed 10% depending on how dose to water in medium is defined

  20. Favorable Preliminary Outcomes for Men With Low- and Intermediate-risk Prostate Cancer Treated With 19-Gy Single-fraction High-dose-rate Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Krauss, Daniel J., E-mail: dkrauss@beaumont.edu [Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States); Ye, Hong [Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States); Martinez, Alvaro A. [21st Century Oncology, Farmington Hills, Michigan (United States); Mitchell, Beth; Sebastian, Evelyn; Limbacher, Amy; Gustafson, Gary S. [Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States)

    2017-01-01

    Purpose: To report the toxicity and preliminary clinical outcomes of a prospective trial evaluating 19-Gy, single-fraction high-dose-rate (HDR) brachytherapy for men with low- and intermediate-risk prostate cancer. Methods and Materials: A total of 63 patients were treated according to an institutional review board-approved prospective study of single-fraction HDR brachytherapy. Eligible patients had tumor stage ≤T2a, prostate-specific antigen level ≤15 ng/mL, and Gleason score ≤7. Patients with a prostate gland volume >50 cm{sup 3} and baseline American Urologic Association symptom score >12 were ineligible. Patients underwent transrectal ultrasound-guided transperineal implantation of the prostate, followed by single-fraction HDR brachytherapy. Treatment was delivered using {sup 192}Ir to a dose of 19 Gy prescribed to the prostate, with no additional margin applied. Results: Of the 63 patients, 58 had data available for analysis. Five patients had withdrawn consent during the follow-up period. The median follow-up period was 2.9 years (range 0.3-5.2). The median age was 61.4 years. The median gland volume at treatment was 34.8 cm{sup 3}. Of the 58 patients, 91% had T1 disease, 71% had Gleason score ≤6 (29% with Gleason score 7), and the median pretreatment prostate-specific antigen level was 5.1 ng/mL. The acute and chronic grade 2 genitourinary toxicity incidence was 12.1% and 10.3%, respectively. No grade 3 urinary toxicity occurred. No patients experienced acute rectal toxicity grade ≥2, and 2 experienced grade ≥2 chronic gastrointestinal toxicity. Three patients experienced biochemical failure, yielding a 3-year cumulative incidence estimate of 6.8%. Conclusions: Single-fraction HDR brachytherapy is well-tolerated, with favorable preliminary biochemical and clinical disease control rates.

  1. Comparison of absorbed dose in the cervix carcinoma therapy by brachytherapy of high dose rate using the conventional planning and Monte Carlo simulation

    International Nuclear Information System (INIS)

    Silva, Aneli Oliveira da

    2010-01-01

    This study aims to compare the doses received for patients submitted to brachytherapy High Dose Rate (HDR) brachytherapy, a method of treatment of the cervix carcinoma, performed in the planning system PLATO BPS with the doses obtained by Monte Carlo simulation using the radiation transport code MCNP 5 and one female anthropomorphic phantom based on voxel, the FAX. The implementation of HDR brachytherapy treatment for the cervix carcinoma consists of the insertion of an intrauterine probe and an intravaginal probe (ring or ovoid) and then two radiographs are obtained, anteroposterior (AP) and lateral (LAT) to confirm the position of the applicators in the patient and to allow the treatment planning and the determination of the absorbed dose at points of interest: rectum, bladder, sigmoid and point A, which corresponds anatomically to the crossings of the uterine arteries with ureters The absorbed doses obtained with the code MCNP 5, with the exception of the absorbed dose in the rectum and sigmoid for the simulation considering a point source of 192 Ir, are lower than the absorbed doses from PLATO BPS calculations because the MCNP 5 considers the chemical compositions and densities of FAX body, not considering the medium as water. When considering the Monte Carlo simulation for a source with dimensions equal to that used in the brachytherapy irradiator used in this study, the values of calculated absorbed dose to the bladder, to the rectum, to the right point A and to the left point A were respectively lower than those determined by the treatment planning system in 33.29, 5.01, 22.93 and 19.04%. These values are almost all larger than the maximum acceptable deviation between patient planned and administered doses (5 %). With regard to the rectum and bladder, which are organs that must be protected, the present results are in favor of the radiological protection of patients. The point A, that is on the isodose of 100%, used to tumor treatment, the results indicate

  2. The feasibility study and characterization of a two-dimensional diode array in “magic phantom†for high dose rate brachytherapy quality assurance

    International Nuclear Information System (INIS)

    Espinoza, A.; Beeksma, B.; Petasecca, M.; Fuduli, I.; Porumb, C.; Cutajar, D.; Lerch, M. L. F.; Rosenfeld, A. B.; Corde, S.; Jackson, M.

    2013-01-01

    Purpose: High dose rate (HDR) brachytherapy is a radiation treatment technique capable of delivering large dose rates to the tumor. Radiation is delivered using remote afterloaders to drive highly active sources (commonly 192 Ir with an air KERMA strength range between 20 000 and 40 000 U, where 1 U = 1 μGy m 2 /h in air) through applicators directly into the patient's prescribed region of treatment. Due to the obvious ramifications of incorrect treatment while using such an active source, it is essential that there are methods for quality assurance (QA) that can directly and accurately verify the treatment plan and the functionality of the remote afterloader. This paper describes the feasibility study of a QA system for HDR brachytherapy using a phantom based two-dimensional 11 × 11 epitaxial diode array, named “magic phantom.â€Methods: The HDR brachytherapy treatment plan is translated to the phantom with two rows of 10 (20 in total) HDR source flexible catheters, arranged above and below the diode array “magic plate†(MP). Four-dimensional source tracking in each catheter is based upon a developed fast iterative algorithm, utilizing the response of the diodes in close proximity to the 192 Ir source, sampled at 100 ms intervals by a fast data acquisition (DAQ) system. Using a 192 Ir source in a solid water phantom, the angular response of the developed epitaxial diodes utilized in the MP and also the variation of the MP response as a function of the source-to-detector distance (SDD) were investigated. These response data are then used by an iterative algorithm for source dwelling position determination. A measurement of the average transit speed between dwell positions was performed using the diodes and a fast DAQ.Results: The angular response of the epitaxial diode showed a variation of 15% within 360°, with two flat regions above and below the detector face with less than 5% variation. For SDD distances of between 5 and 30 mm the relative response of

  3. Using the computed tomography in comparison to the orthogonal radiography based treatment planning in high dose rate (HDR) brachytherapy in cervical uteri cancer patients; a single institution feasibility study.

    Science.gov (United States)

    Bahadur, Yasir A; El-Sayed, Mohamed E; El-Taher, Zeinab H; Zaza, Khaled O; Moftah, Belal A; Hassouna, Ashraf H; Ghassal, Noor M

    2008-03-01

    Brachytherapy is an integral part in the treatment of cervical uteri cancer patients. Orthogonal treatment planning is the standard mode of calculation based on reference points. Introduction of the innovative 3-D computer based treatment planning allows accurate calculation based on volumetric information as regards the target volume and organs at risk (OAR). Also provide dose volume histogram (DVH) for proper estimation of the dose in relation to the volume. To correlate and compare the information obtained from the two approaches for high dose rate brachytherapy of cervical uteri cancer; the orthogonal conventional method and the computerized tomography (CT) three dimensions (3D) based calculation method in relation to the target and organ at risk (OAR). From 6 patients of cervical uteri cancer, 21 applications with orthogonal planning using the Brachy Vision treatment planning system version 7.3.10 were performed. In 10 applications; comparison between orthogonal and CT based planning was done. In orthogonal planning; the dose to point A, rectum and bladder were defined according to the American Brachytherapy Society (ABS) recommendation. From the CT based planning the target volume and dose volume histogram lpar;DVH) were calculated for the clinical target volume (CTV), rectum and bladder. From these two sets, information was obtained and compared and mean values were derived. For dose prescription at point A, an average of 63.5% of CTV received the prescribed dose. The mean ICRU dose to the bladder point is 2.9 Gy+/-1.2 SD (Standard Deviation) and 17% of the bladder volume derived from CT was encompassed by 2.9 Gy isodose line. The mean ICRU dose at the rectum point is 3.4 Gy+/-1.2 SD and 21% of the rectum volume from CT was encompassed by 3.4 Gy isodose line. The maximum dose to the rectum and the bladder derived from the CT and compared to the maximal dose at ICRU is 1.7 and 2.8 times higher than the orthogonal reference points; with the corresponding p

  4. Using the Computed Tomography in Comparison to the Orthogonal Radiography Based Treatment Planning in High dose Rate (HDR) Brachytherapy in Cervical Uteri Cancer Patients; A Single Institution Feasibility Study

    International Nuclear Information System (INIS)

    BAHADUR, Y.A.; EL-SAYED, M.E.; HASSOUNA, A.H.; EL-TAHER, Z.H.; GHASSAL, N.M.; ZAZA, Kh.O.M.D.; OFTAH, B.A.

    2008-01-01

    Brachytherapy is an integral part in the treatment of cervical uteri cancer patients. Orthogonal treatment planning is the standard mode of calculation based on reference points. Introduction of the innovative 3-D computer based treatment planning allows accurate calculation based on volumetric information as regards the target volume and organs at risk (OAR). Also provide dose volume histogram (DVH) for proper estimation of the dose in relation to the volume. Aim: To correlate and compare the information obtained from the two approaches for high dose rate brachytherapy of cervical uteri cancer; the orthogonal conventional method and the computerized tomography (CT) three dimensions (3D) based calculation method in relation to the target and organ at risk (OAR). Methods: From 6 patients of cervical uteri cancer, 21 applications with orthogonal planning using the Brachy Vision treatment planning system version 7.3.10 were performed. In 10 applications; comparison between orthogonal and CT based planning was done. In orthogonal planning; the dose to point A, rectum and bladder were defined according to the American Brachytherapy Society (ABS) recommendation. From the CT based planning the target volume and dose volume histogram (DVH) were calculated for the clinical target volume (CTV), rectum and bladder. From these two sets, information was obtained and compared and mean values were derived. Results: For dose prescription at point A, an average of 63.5% of CTV received the prescribed dose. The mean ICRU dose to the bladder point is 2.9 Gy±l .2 SD (Standard Deviation) and 17% of the bladder volume derived from CT was encompassed by 2.9 Gy isodose line. The mean ICRU dose at the rectum point is 3.4 Gy±1.2 SD and 21% of the rectum volume from CT was encompassed by 3.4 Gy isodose line. The maximum dose to the rectum and the bladder derived from the CT and compared to the maximal dose at ICRU is 1.7 and 2.8 times higher than the orthogonal reference points; with the

  5. Design and testing of a phantom and instrumented gynecological applicator based on GaN dosimeter for use in high dose rate brachytherapy quality assurance

    Energy Technology Data Exchange (ETDEWEB)

    Guiral, P.; Ribouton, J.; Jalade, P. [Service de Physique Médicale et Radioprotection, Centre Hospitalier Lyon Sud, Pierre-Bénite F-69495 (France); Wang, R.; Galvan, J.-M.; Lu, G.-N.; Pittet, P., E-mail: patrick.pittet@univ-lyon1.fr [Institut des Nanotechnologies de Lyon, University Lyon, Université Claude Bernard Lyon 1, CNRS, INL UMR5270, F-69622 Villeurbanne (France); Rivoire, A.; Gindraux, L. [DOSILAB, 66 Boulevard Niels Bohr, Villeurbanne F-69100 (France)

    2016-09-15

    Purpose: High dose rate brachytherapy (HDR-BT) is widely used to treat gynecologic, anal, prostate, head, neck, and breast cancers. These treatments are typically administered in large dose per fraction (>5 Gy) and with high-gradient-dose-distributions, with serious consequences in case of a treatment delivery error (e.g., on dwell position and dwell time). Thus, quality assurance (QA) or quality control (QC) should be systematically and independently implemented. This paper describes the design and testing of a phantom and an instrumented gynecological applicator for pretreatment QA and in vivo QC, respectively. Methods: The authors have designed a HDR-BT phantom equipped with four GaN-based dosimeters. The authors have also instrumented a commercial multichannel HDR-BT gynecological applicator by rigid incorporation of four GaN-based dosimeters in four channels. Specific methods based on the four GaN dosimeter responses are proposed for accurate determination of dwell time and dwell position inside phantom or applicator. The phantom and the applicator have been tested for HDR-BT QA in routine over two different periods: 29 and 15 days, respectively. Measurements in dwell position and time are compared to the treatment plan. A modified position–time gamma index is used to monitor the quality of treatment delivery. Results: The HDR-BT phantom and the instrumented applicator have been used to determine more than 900 dwell positions over the different testing periods. The errors between the planned and measured dwell positions are 0.11 ± 0.70 mm (1σ) and 0.01 ± 0.42 mm (1σ), with the phantom and the applicator, respectively. The dwell time errors for these positions do not exhibit significant bias, with a standard deviation of less than 100 ms for both systems. The modified position–time gamma index sets a threshold, determining whether the treatment run passes or fails. The error detectability of their systems has been evaluated through tests on

  6. A JPEG backward-compatible HDR image compression

    Science.gov (United States)

    Korshunov, Pavel; Ebrahimi, Touradj

    2012-10-01

    High Dynamic Range (HDR) imaging is expected to become one of the technologies that could shape next generation of consumer digital photography. Manufacturers are rolling out cameras and displays capable of capturing and rendering HDR images. The popularity and full public adoption of HDR content is however hindered by the lack of standards in evaluation of quality, file formats, and compression, as well as large legacy base of Low Dynamic Range (LDR) displays that are unable to render HDR. To facilitate wide spread of HDR usage, the backward compatibility of HDR technology with commonly used legacy image storage, rendering, and compression is necessary. Although many tone-mapping algorithms were developed for generating viewable LDR images from HDR content, there is no consensus on which algorithm to use and under which conditions. This paper, via a series of subjective evaluations, demonstrates the dependency of perceived quality of the tone-mapped LDR images on environmental parameters and image content. Based on the results of subjective tests, it proposes to extend JPEG file format, as the most popular image format, in a backward compatible manner to also deal with HDR pictures. To this end, the paper provides an architecture to achieve such backward compatibility with JPEG and demonstrates efficiency of a simple implementation of this framework when compared to the state of the art HDR image compression.

  7. Real-Time Verification of a High-Dose-Rate Iridium 192 Source Position Using a Modified C-Arm Fluoroscope

    Energy Technology Data Exchange (ETDEWEB)

    Nose, Takayuki, E-mail: nose-takayuki@nms.ac.jp [Department of Radiation Oncology, Nippon Medical School Tamanagayama Hospital, Tama (Japan); Chatani, Masashi [Department of Radiation Oncology, Osaka Rosai Hospital, Sakai (Japan); Otani, Yuki [Department of Radiology, Kaizuka City Hospital, Kaizuka (Japan); Teshima, Teruki [Department of Radiation Oncology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka (Japan); Kumita, Shinichirou [Department of Radiology, Nippon Medical School Hospital, Tokyo (Japan)

    2017-03-15

    Purpose: High-dose-rate (HDR) brachytherapy misdeliveries can occur at any institution, and they can cause disastrous results. Even a patient's death has been reported. Misdeliveries could be avoided with real-time verification methods. In 1996, we developed a modified C-arm fluoroscopic verification of an HDR Iridium 192 source position prevent these misdeliveries. This method provided excellent image quality sufficient to detect errors, and it has been in clinical use at our institutions for 20 years. The purpose of the current study is to introduce the mechanisms and validity of our straightforward C-arm fluoroscopic verification method. Methods and Materials: Conventional X-ray fluoroscopic images are degraded by spurious signals and quantum noise from Iridium 192 photons, which make source verification impractical. To improve image quality, we quadrupled the C-arm fluoroscopic X-ray dose per pulse. The pulse rate was reduced by a factor of 4 to keep the average exposure compliant with Japanese medical regulations. The images were then displayed with quarter-frame rates. Results: Sufficient quality was obtained to enable observation of the source position relative to both the applicators and the anatomy. With this method, 2 errors were detected among 2031 treatment sessions for 370 patients within a 6-year period. Conclusions: With the use of a modified C-arm fluoroscopic verification method, treatment errors that were otherwise overlooked were detected in real time. This method should be given consideration for widespread use.

  8. Relative effect of dose-rate values and fractionation on late responding tissues and tumours

    International Nuclear Information System (INIS)

    Malgieri, F.

    1995-01-01

    There are currently available different facilities for radiotherapy also with regard to the dose-rate values (in the ranges LDR - MDR - HDR), sometimes used alternatively or subsequently for the same tumour. We have set up a 'unitary' L-Q model, based on Liversage's and Dale's works, that explicitly include also the dose-rate value and a correction factor of the β parameter depending on the sublethal damage repair time constant, on the length of time of each irradiation and on the time interval between following irradiation for to realize the effect of the incomplete repair when the time interval is short as, for example, in the PLDR. This 'unitary' L-Q model is, of course, usable in the same way both for external beam therapy and for curietherapy and make possible to compute and compare, for each kind of tumour and normal tissue, the relative effect of the different available modality of radiotherapy also with regard to the dose-rate. We show and discuss the resulting relationships of the ratio BED 'late'/BED tumour changing the time-dose parameters and the values of the biological characteristic parameters T p , α/β and μ, for defined size of tumour control and different value of the doserate

  9. Dwell time modulation restrictions do not necessarily improve treatment plan quality for prostate HDR brachytherapy

    International Nuclear Information System (INIS)

    Balvert, Marleen; Gorissen, Bram L; Den Hertog, Dick; Hoffmann, Aswin L

    2015-01-01

    Inverse planning algorithms for dwell time optimisation in interstitial high-dose-rate (HDR) brachytherapy may produce solutions with large dwell time variations within catheters, which may result in undesirable selective high-dose subvolumes. Extending the dwell time optimisation model with a dwell time modulation restriction (DTMR) that limits dwell time differences between neighboring dwell positions has been suggested to eliminate this problem. DTMRs may additionally reduce the sensitivity for uncertainties in dwell positions that inevitably result from catheter reconstruction errors and afterloader source positioning inaccuracies. This study quantifies the reduction of high-dose subvolumes and the robustness against these uncertainties by applying a DTMR to template-based prostate HDR brachytherapy implants. Three different DTMRs were consecutively applied to a linear dose-based penalty model (LD) and a dose-volume based model (LDV), both obtained from literature. The models were solved with DTMR levels ranging from no restriction to uniform dwell times within catheters in discrete steps. Uncertainties were simulated on clinical cases using in-house developed software, and dose-volume metrics were calculated in each simulation. For the assessment of high-dose subvolumes, the dose homogeneity index (DHI) and the contiguous dose volume histogram were analysed. Robustness was measured by the improvement of the lowest D 90% of the planning target volume (PTV) observed in the simulations. For (LD), a DTMR yields an increase in DHI of approximately 30% and reduces the size of the largest high-dose volume by 2–5 cc. However, this comes at a cost of a reduction in D 90% of the PTV of 10%, which often implies that it drops below the desired minimum of 100%. For (LDV), none of the DTMRs were able to improve high-dose volume measures. DTMRs were not capable of improving robustness of PTV D 90% against uncertainty in dwell positions for both models. (paper)

  10. Six fractions per week of external beam radiotherapy and high-dose-rate brachytherapy for carcinoma of the uterine cervix: A phase I/II study

    International Nuclear Information System (INIS)

    Yoon, Sang Min; Huh, Seung Jae; Park, Won; Lee, Jeung Eun; Park, Young Je; Nam, Hee Rim; Lim, Do Hoon; Ahn, Yong Chan

    2006-01-01

    Purpose: This study evaluated the treatment results of external beam radiotherapy administered in six fractions per week and high-dose-rate (HDR) brachytherapy for the treatment of cervical cancer. Methods and Materials: From July 2000 to July 2003, 43 patients were enrolled in this study. The patients received 45 Gy from a 10-MV photon beam using four-field box or anterior-posterior beams. Parametrial regions and the pelvic side walls were boosted with up to 50.4 Gy using a midline block. The daily fraction dose was 1.8 Gy administered in six-weekly fractions, from Monday to Saturday. HDR brachytherapy was also delivered at doses of 24 Gy to point A in six fractions twice a week. The median follow-up time was 37 months (range, 9-60 months). Results: The median overall treatment time was 51 days for all patients (range, 44-62 days). Thirty-four patients (79.1%) achieved complete remission and 8 (18.6%) achieved partial remission after radiotherapy. Locoregional recurrence occurred in 5 patients (11.6%), and a distant metastasis was encountered in 6 patients (13.9%). The 3-year overall survival, locoregional, and distant metastasis-free survival rates were 74.7%, 87.8%, and 84.7%, respectively. Grade 2 and 3 late rectal complications were encountered in 3 (6.5%) and 1 (2.2%), respectively. There were no Grade 3 late bladder complications. Conclusions: Six fractions per week of external beam radiotherapy and HDR brachytherapy is an effective treatment for patients with a carcinoma of the uterine cervix and can be used as a possible alternative to concomitant chemoradiotherapy in elderly patients or in patients with co-morbidity

  11. Dosimetric Evaluation of High-Dose-Rate Interstitial Brachytherapy Boost Treatments for Localized Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Froehlich, Georgina [Semmelweis Univ., Budapest (Hungary); Dept. of Radiotherapy, National Inst. of Oncology, Budapest (Hungary); Agoston, Peter; Loevey, Jozsef; Somogyi, Andras; Fodor, Janos; Polgar, Csaba; Major, Tibor [Dept. of Radiotherapy, National Inst. of Oncology, Budapest (Hungary)

    2010-07-15

    Purpose: to quantitatively evaluate the dose distributions of high-dose-rate (HDR) prostate implants regarding target coverage, dose homogeneity, and dose to organs at risk. Material and methods: treatment plans of 174 implants were evaluated using cumulative dose-volume histograms (DVHs). The planning was based on transrectal ultrasound (US) imaging, and the prescribed dose (100%) was 10 Gy. The tolerance doses to rectum and urethra were 80% and 120%, respectively. Dose-volume parameters for target (V90, V100, V150, V200, D90, D{sub min}) and quality indices (DNR [dose nonuniformity ratio], DHI [dose homogeneity index], CI [coverage index], COIN [conformal index]) were calculated. Maximum dose in reference points of rectum (D{sub r}) and urethra (D{sub u}), dose to volume of 2 cm{sup 3} of the rectum (D{sub 2ccm}), and 0.1 cm{sup 3} and 1% of the urethra (D{sub 0.1ccm} and D1) were determined. Nonparametric correlation analysis was performed between these parameters. Results: the median number of needles was 16, the mean prostate volume (V{sub p}) was 27.1 cm{sup 3}. The mean V90, V100, V150, and V200 were 90%, 97%, 39% and 13%, respectively. The mean D90 was 109%, and the D{sub min} was 87%. The mean doses in rectum and urethra reference points were 75% and 119%, respectively. The mean volumetric doses were D{sub 2ccm} = 49% for the rectum, D{sub 0.1ccm} = 126%, and D1 = 140% for the urethra. The mean DNR was 0.37, while the DHI was 0.60. The mean COIN was 0.66. The Spearman rank order correlation coefficients for volume doses to rectum and urethra were R(D{sub r}, D{sub 2ccm}) = 0.69, R(D{sub u}, D{sub 0.1ccm}) = 0.64, R(D{sub u}, D1) = 0.23. Conclusion: US-based treatment plans for HDR prostate implants based on the real positions of catheters provided acceptable dose distributions. In the majority of the cases, the doses to urethra and rectum were kept below the defined tolerance levels. For rectum, the dose in reference points correlated well with dose

  12. Bystander effects of exposure to low-dose-rate 125I seeds on human lung cancers cells in vitro

    International Nuclear Information System (INIS)

    Jia Rongfei; Chen Honghong; Yu Lei; Zhao Meijia; Shao Chunlin; Cheng Wenying

    2007-01-01

    The bystander effects induced by continuous low-dose-rate (LDR) 125 I seeds radiation on damage of human lung cancer cells were investigated. Human adenocarcinoma cell line A549 and human small cell lung cancer cell line NCI-H446, which have different sensitivities to high-dose rate (HDR) external irradiation, were exposed directly to 125 I seeds in vitro and co-cultured with unirradiated cells for 24 h. Using cytokinesis-blocking micronucleus method and γ H2AX fluorescence immunoassay, bystander effects induced by 2Gy and 4Gy 125 I seed irradiation on micronucleus formation and DNA double-strand breaks (DSBs) of human lung cancer cells were detected and evaluated. The results showed that irradiation with 125 I seeds can induce medium-mediated bystander effects in A549 cells and NCI-H446 cells, exhibiting that both micronuclei formation and γ H2AX focus formation in bystander cells were increased significantly compared with non-irradiated cells. The extent of DNA damage induced by bystander effects was correlated with accumulated radiation dose and radiosensitive of tumor cells. NCI-H446 cells that were sensitive to HDR γ irradiation were more sensitive to continuous LDR irradiation and bystander effects than A549. However, a comparison between the bystander effects and direct effects elicits the intensity of bystander responses of A549 cells was higher than that of NCI-H446 cells. A dose-related reduction in bystander responses was observed both in A549 cells and NCI-H446 cells, suggesting that the signaling factors involved in the bystander signaling pathways may decrease with the increase of cell damages. (authors)

  13. Clinically evident fat necrosis in women treated with high-dose-rate brachytherapy alone for early-stage breast cancer

    International Nuclear Information System (INIS)

    Wazer, David E.; Lowther, David; Boyle, Teresa; Ulin, Kenneth; Neuschatz, Andrew; Ruthazer, Robin; DiPetrillo, Thomas A.

    2001-01-01

    Purpose: To investigate the incidence of and variables associated with clinically evident fat necrosis in women treated on a protocol of high-dose-rate (HDR) brachytherapy alone without external-beam whole-breast irradiation for early-stage breast carcinoma. Methods and Materials: From 6/1997 until 8/1999, 30 women diagnosed with Stage I or II breast carcinoma underwent surgical excision and postoperative irradiation via HDR brachytherapy implant as part of a multi-institutional clinical Phase I/II protocol. Patients eligible included those with T1, T2, N0, N1 (≤3 nodes positive), M0 tumors of nonlobular histology with negative surgical margins, no extracapsular lymph-node extension, and a negative postexcision mammogram. Brachytherapy catheters were placed at the initial excision, re-excision, or at the time of axillary sampling. Direct visualization, surgical clips, ultrasound, or CT scans assisted in delineating the target volume defined as the excision cavity plus 2-cm margin. High activity 192 Ir (3-10 Ci) was used to deliver 340 cGy per fraction, 2 fractions per day, for 5 consecutive days to a total dose of 34 Gy to the target volume. Source position and dwell times were calculated using standard volume optimization techniques. Dosimetric analyses were performed with three-dimensional postimplant dose and volume reconstructions. The median follow-up of all patients was 24 months (range, 12-36 months). Results: Eight patients (crude incidence of 27%) developed clinically evident fat necrosis postimplant in the treated breast. Fat necrosis was determined by clinical presentation including pain and swelling in the treated volume, computed tomography, and/or biopsy. All symptomatic patients (7 of 8 cases) were successfully treated with 3 to 12 months of conservative management. Continuous variables that were found to be associated significantly with fat necrosis included the number of source dwell positions (p=0.04), and the volume of tissue which received

  14. SU-E-T-574: Fessiblity of Using the Calypso System for HDR Interstitial Catheter Reconstruction

    International Nuclear Information System (INIS)

    Li, J S; Ma, C

    2014-01-01

    Purpose: It is always a challenge to reconstruct the interstitial catheter for high dose rate (HDR) brachytherapy on patient CT or MR images. This work aims to investigate the feasibility of using the Calypso system (Varian Medical, CA) for HDR catheter reconstruction utilizing its accuracy on tracking the electromagnetic transponder location. Methods: Experiment was done with a phantom that has a HDR interstitial catheter embedded inside. CT scan with a slice thickness of 1.25 mm was taken for this phantom with two Calypso beacon transponders in the catheter. The two transponders were connected with a wire. The Calypso system was used to record the beacon transponders’ location in real time when they were gently pulled out with the wire. The initial locations of the beacon transponders were used for registration with the CT image and the detected transponder locations were used for the catheter path reconstruction. The reconstructed catheter path was validated on the CT image. Results: The HDR interstitial catheter was successfully reconstructed based on the transponders’ coordinates recorded by the Calypso system in real time when the transponders were pulled in the catheter. After registration with the CT image, the shape and location of the reconstructed catheter are evaluated against the CT image and the result shows an accuracy of 2 mm anywhere in the Calypso detectable region which is within a 10 cm X 10 cm X 10 cm cubic box for the current system. Conclusion: It is feasible to use the Calypso system for HDR interstitial catheter reconstruction. The obstacle for its clinical usage is the size of the beacon transponder whose diameter is bigger than most of the interstitial catheters used in clinic. Developing smaller transponders and supporting software and hardware for this application is necessary before it can be adopted for clinical use

  15. Clinical implementation of a new HDR brachytherapy device for partial breast irradiation

    International Nuclear Information System (INIS)

    Scanderbeg, Daniel J.; Yashar, Catheryn; Rice, Roger; Pawlicki, Todd

    2009-01-01

    Purpose: To present the clinical implementation of a new HDR device for partial breast irradiation, the Strut-Adjusted Volume Implant (SAVI), at the University of California, San Diego. Methods and materials: The SAVI device has multiple peripheral struts that can be differentially loaded with the HDR source. Planning criteria used for evaluation of the treatment plans included the following dose volume histogram (DVH) criteria: V90 >90%, V150 <50 cc and V200 <20 cc. Results: SAVI has been used on 20 patients to date at UC San Diego. In each case, the dose was modulated according to patient-specific anatomy to cover the tumor bed, while sparing normal tissues. The dosimetric data show that we can achieve greater than 90% coverage with respect to V90 (median of 95.3%) and also keep a low V150 and V200 dose at 24.5 and 11.2 cc, respectively. Complete treatment can be done within a 30-min time slot, which includes implant verification, setup, and irradiation time as well as wound dressing. Conclusion: SAVI has been implemented at UC San Diego for accelerated partial breast irradiation with excellent tumor bed conformance and minimal normal tissue exposure. Patient positioning is the key to identifying any inter-fraction device motion. Device asymmetry or tissue conformance has been shown to resolve itself 24 h after the device implantation. The device can be implemented into an existing HDR program with minimal effort

  16. High-Dose-Rate Brachytherapy as a Monotherapy for Favorable-Risk Prostate Cancer: A Phase II Trial

    International Nuclear Information System (INIS)

    Barkati, Maroie; Williams, Scott G.; Foroudi, Farshad; Tai, Keen Hun; Chander, Sarat; Dyk, Sylvia van; See, Andrew; Duchesne, Gillian M.

    2012-01-01

    Purpose: There are multiple treatment options for favorable-risk prostate cancer. High-dose-rate (HDR) brachytherapy as a monotherapy is appealing, but its use is still investigational. A Phase II trial was undertaken to explore the value of such treatment in low-to-intermediate risk prostate cancer. Methods and Materials: This was a single-institution, prospective study. Eligible patients had low-risk prostate cancer features but also Gleason scores of 7 (51% of patients) and stage T2b to T2c cancer. Treatment with HDR brachytherapy with a single implant was administered over 2 days. One of four fractionation schedules was used in a dose escalation study design: 3 fractions of 10, 10.5, 11, or 11.5 Gy. Patients were assessed with the Common Terminology Criteria for Adverse Events version 2.0 for urinary toxicity, the Radiation Therapy Oncology Group/European Organization for Research and Treatment of Cancer scoring schema for rectal toxicity, and the Expanded Prostate Cancer Index Composite (EPIC) questionnaire to measure patient-reported health-related quality of life. Biochemical failure was defined as a prostate-specific antigen (PSA) nadir plus 2 ng/ml. Results: Between 2003 and 2008, 79 patients were enrolled. With a median follow-up of 39.5 months, biochemical relapse occurred in 7 patients. Three- and 5-year actuarial biochemical control rates were 88.4% (95% confidence interval [CI], 78.0-96.2%) and 85.1% (95% CI, 72.5-94.5%), respectively. Acute grade 3 urinary toxicity was seen in only 1 patient. There was no instance of acute grade 3 rectal toxicity. Rates of late grade 3 rectal toxicity, dysuria, hematuria, urinary retention, and urinary incontinence were 0%, 10.3%, 1.3%, 9.0%, and 0%, respectively. No grade 4 or greater toxicity was recorded. Among the four (urinary, bowel, sexual, and hormonal) domains assessed with the EPIC questionnaire, only the sexual domain did not recover with time. Conclusions: HDR brachytherapy as a monotherapy for favorable

  17. High-Dose-Rate Brachytherapy as a Monotherapy for Favorable-Risk Prostate Cancer: A Phase II Trial

    Energy Technology Data Exchange (ETDEWEB)

    Barkati, Maroie [Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne (Australia); Williams, Scott G., E-mail: scott.williams@petermac.org [Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne (Australia); Department of Pathology, University of Melbourne, Melbourne (Australia); Foroudi, Farshad; Tai, Keen Hun; Chander, Sarat [Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne (Australia); Department of Pathology, University of Melbourne, Melbourne (Australia); Dyk, Sylvia van [Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne (Australia); See, Andrew [Ballarat Austin Radiation Oncology Centre, Ballarat (Australia); Duchesne, Gillian M. [Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne (Australia); Department of Pathology, University of Melbourne, Melbourne (Australia)

    2012-04-01

    Purpose: There are multiple treatment options for favorable-risk prostate cancer. High-dose-rate (HDR) brachytherapy as a monotherapy is appealing, but its use is still investigational. A Phase II trial was undertaken to explore the value of such treatment in low-to-intermediate risk prostate cancer. Methods and Materials: This was a single-institution, prospective study. Eligible patients had low-risk prostate cancer features but also Gleason scores of 7 (51% of patients) and stage T2b to T2c cancer. Treatment with HDR brachytherapy with a single implant was administered over 2 days. One of four fractionation schedules was used in a dose escalation study design: 3 fractions of 10, 10.5, 11, or 11.5 Gy. Patients were assessed with the Common Terminology Criteria for Adverse Events version 2.0 for urinary toxicity, the Radiation Therapy Oncology Group/European Organization for Research and Treatment of Cancer scoring schema for rectal toxicity, and the Expanded Prostate Cancer Index Composite (EPIC) questionnaire to measure patient-reported health-related quality of life. Biochemical failure was defined as a prostate-specific antigen (PSA) nadir plus 2 ng/ml. Results: Between 2003 and 2008, 79 patients were enrolled. With a median follow-up of 39.5 months, biochemical relapse occurred in 7 patients. Three- and 5-year actuarial biochemical control rates were 88.4% (95% confidence interval [CI], 78.0-96.2%) and 85.1% (95% CI, 72.5-94.5%), respectively. Acute grade 3 urinary toxicity was seen in only 1 patient. There was no instance of acute grade 3 rectal toxicity. Rates of late grade 3 rectal toxicity, dysuria, hematuria, urinary retention, and urinary incontinence were 0%, 10.3%, 1.3%, 9.0%, and 0%, respectively. No grade 4 or greater toxicity was recorded. Among the four (urinary, bowel, sexual, and hormonal) domains assessed with the EPIC questionnaire, only the sexual domain did not recover with time. Conclusions: HDR brachytherapy as a monotherapy for favorable

  18. Application of RADPOS in Vivo Dosimetry for QA of High Dose Rate Brachytherapy

    DEFF Research Database (Denmark)

    Cherpak, A.; Kertzscher Schwencke, Gustavo Adolfo Vladimir; Cygler, J.

    2012-01-01

    cancer, where high dose gradients and movement of the prostate gland can present unique in vivo dosimetry challenges. Financial and technical support has been received from Best Medical Canada and Ascension Technology Corporation. © 2012 American Association of Physicists in Medicine......Purpose: The RADPOS in vivo dosimetry system combines an electromagnetic positioning sensor with MOSFET dosimetry, allowing for simultaneous online measurements of dose and spatial position. In this work, we assess the potential use of RADPOS for measurements of motion and dose during prostate HDR...

  19. WE-DE-201-11: Sensitivity and Specificity of Verification Methods Based On Total Reference Air Kerma (TRAK) Or On User Provided Dose Points for Graphically Planned Skin HDR Brachytherapy

    International Nuclear Information System (INIS)

    Damato, A; Devlin, P; Bhagwat, M; Buzurovic, I; Hansen, J; O’Farrell, D; Cormack, R

    2016-01-01

    Purpose: To investigate the sensitivity and specificity of a novel verification methodology for image-guided skin HDR brachytherapy plans using a TRAK-based reasonableness test, compared to a typical manual verification methodology. Methods: Two methodologies were used to flag treatment plans necessitating additional review due to a potential discrepancy of 3 mm between planned dose and clinical target in the skin. Manual verification was used to calculate the discrepancy between the average dose to points positioned at time of planning representative of the prescribed depth and the expected prescription dose. Automatic verification was used to calculate the discrepancy between TRAK of the clinical plan and its expected value, which was calculated using standard plans with varying curvatures, ranging from flat to cylindrically circumferential. A plan was flagged if a discrepancy >10% was observed. Sensitivity and specificity were calculated using as a criteria for true positive that >10% of plan dwells had a distance to prescription dose >1 mm different than prescription depth (3 mm + size of applicator). All HDR image-based skin brachytherapy plans treated at our institution in 2013 were analyzed. Results: 108 surface applicator plans to treat skin of the face, scalp, limbs, feet, hands or abdomen were analyzed. Median number of catheters was 19 (range, 4 to 71) and median number of dwells was 257 (range, 20 to 1100). Sensitivity/specificity were 57%/78% for manual and 70%/89% for automatic verification. Conclusion: A check based on expected TRAK value is feasible for irregularly shaped, image-guided skin HDR brachytherapy. This test yielded higher sensitivity and specificity than a test based on the identification of representative points, and can be implemented with a dedicated calculation code or with pre-calculated lookup tables of ideally shaped, uniform surface applicators.

  20. WE-DE-201-11: Sensitivity and Specificity of Verification Methods Based On Total Reference Air Kerma (TRAK) Or On User Provided Dose Points for Graphically Planned Skin HDR Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Damato, A; Devlin, P; Bhagwat, M; Buzurovic, I; Hansen, J; O’Farrell, D; Cormack, R [Harvard Medical School, Boston, MA (United States)

    2016-06-15

    Purpose: To investigate the sensitivity and specificity of a novel verification methodology for image-guided skin HDR brachytherapy plans using a TRAK-based reasonableness test, compared to a typical manual verification methodology. Methods: Two methodologies were used to flag treatment plans necessitating additional review due to a potential discrepancy of 3 mm between planned dose and clinical target in the skin. Manual verification was used to calculate the discrepancy between the average dose to points positioned at time of planning representative of the prescribed depth and the expected prescription dose. Automatic verification was used to calculate the discrepancy between TRAK of the clinical plan and its expected value, which was calculated using standard plans with varying curvatures, ranging from flat to cylindrically circumferential. A plan was flagged if a discrepancy >10% was observed. Sensitivity and specificity were calculated using as a criteria for true positive that >10% of plan dwells had a distance to prescription dose >1 mm different than prescription depth (3 mm + size of applicator). All HDR image-based skin brachytherapy plans treated at our institution in 2013 were analyzed. Results: 108 surface applicator plans to treat skin of the face, scalp, limbs, feet, hands or abdomen were analyzed. Median number of catheters was 19 (range, 4 to 71) and median number of dwells was 257 (range, 20 to 1100). Sensitivity/specificity were 57%/78% for manual and 70%/89% for automatic verification. Conclusion: A check based on expected TRAK value is feasible for irregularly shaped, image-guided skin HDR brachytherapy. This test yielded higher sensitivity and specificity than a test based on the identification of representative points, and can be implemented with a dedicated calculation code or with pre-calculated lookup tables of ideally shaped, uniform surface applicators.

  1. Medium-dose-rate intracavitary brachytherapy for cervical cancer

    International Nuclear Information System (INIS)

    Tanaka, Eiichi; Isohashi, Fumiaki; Oh, Ryoong-Jin

    2003-01-01

    due to pelvic necrosis without local recurrence 6 years after radiotherapy (Grade 5). Another patient developed perforation of sigmoid colon 9 years after radiotherapy (Grade 4). Minor late complications (Grade 1-2) occurred in 7 patients (9%). MDR-ICRT for cervical cancer can be used as effectively as low-dose-late (LDR) and high-dose-rate (HDR) ICRT. (author)

  2. TU-C-201-00: Clinical Implementation of HDR Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2015-06-15

    Recent use of HDR has increased while planning has become more complex often necessitating 3D image-based planning. While many guidelines for the use of HDR exist, they have not kept pace with the increased complexity of 3D image-based planning. Furthermore, no comprehensive document exists to describe the wide variety of current HDR clinical indications. This educational session aims to summarize existing national and international guidelines for the safe implementation of an HDR program. A summary of HDR afterloaders available on the market and their existing applicators will be provided, with guidance on how to select the best fit for each institution’s needs. Finally, the use of checklists will be discussed as a means to implement a safe and efficient HDR program and as a method by which to verify the quality of an existing HDR program. This session will provide the perspective of expert HDR physicists as well as the perspective of a new HDR user. Learning Objectives: Summarize national and international safety and staffing guidelines for HDR implementation Discuss the process of afterloader and applicator selection for gynecologic, prostate, breast, interstitial, surface treatments Learn about the use of an audit checklist tool to measure of quality control of a new or existing HDR program Describe the evolving use of checklists within an HDR program.

  3. Computed Tomography–Planned High-Dose-Rate Brachytherapy for Treating Uterine Cervical Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Zolciak-Siwinska, Agnieszka, E-mail: agnieszka.zolciak@wp.pl [Department of Brachytherapy, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Gruszczynska, Ewelina; Bijok, Michal [Department of Medical Physics, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Jonska-Gmyrek, Joanna [Department of Teleradiotherapy, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Dabkowski, Mateusz [Department of Brachytherapy, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Staniaszek, Jagna [Department of Teleradiotherapy, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Michalski, Wojciech [Department of Clinical Trials and Biostatistics, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland); Kowalczyk, Adam; Milanowska, Katarzyna [Department of Medical Physics, The Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw (Poland)

    2016-09-01

    Purpose: To evaluate the long-term results of computed tomography (CT)–planned high-dose-rate (HDR) brachytherapy (BT) for treating cervical cancer patients. Methods and Materials: CT-planned HDR BT was performed according to the adapted Group European de Curietherapie-European Society for Therapeutic Radiology and Oncology (GEC-ESTRO) recommendations in 216 consecutive patients with locally advanced cervical cancer, International Federation of Gynecology and Obstetrics (FIGO) stage IB to IVA, who were treated with conformal external beam radiation therapy and concomitant chemotherapy. We analyzed outcomes and late side effects evaluated according to the Radiation Therapy Oncology Group/European Organization for Research and Treatment of Cancer and Subjective, Objective, Management, Analysis evaluation scoring system and compared them with the results from a historical group. Results: The median age was 56 years (range, 32-83 years). The median follow-up time for living patients was 52 months (range 37-63 months). The 5-year cumulative incidence function for the local recurrence rate for patients with FIGO II and III was 5.5% and 20%, respectively (P=.001). The 5-year rates of overall survival (OS) and disease-free survival (DFS) were 66.4% and 58.5%, respectively. The relative risk of failure for OS and DFS for FIGO III in relation to FIGO II was 2.24 (P=.003) and 2.6 (P=.000) and for lymph node enlargement was 2.3 (P=.002) and 2 (P=.006), respectively. In 2 patients, rectovaginal fistula occurred, and in 1 patient, vesicovaginal fistula occurred without local progression. Comparison of late adverse effects in patients treated according to the GEC-ESTRO recommendations and in the historical group revealed a reduction in fistula formation of 59% and also a reduction in rectal grade 3 to 4 late toxicity of >59%. Conclusions: This is the largest report with mature data of CT-planned BT HDR for the treatment of cervical cancer with good local control and

  4. Salvage prostate HDR brachytherapy combined with interstitial hyperthermia for local recurrence after radiation therapy failure

    Energy Technology Data Exchange (ETDEWEB)

    Kukielka, A.M.; Hetnal, M.; Dabrowski, T.; Walasek, T.; Brandys, P.; Reinfuss, M. [Centre of Oncology, M. Sklodowska - Curie Institute, Krakow Branch, Department of Radiotherapy, Krakow (Poland); Nahajowski, D.; Kudzia, R.; Dybek, D. [Centre of Oncology, M. Sklodowska - Curie Institute, Krakow Branch, Department of Medical Physics, Department of Radiotherapy, Krakow (Poland)

    2014-02-15

    The aim of the present retrospective study is to evaluate toxicity and early clinical outcomes of interstitial hyperthermia (IHT) combined with high-dose rate (HDR) brachytherapy as a salvage treatment in patients with biopsy-confirmed local recurrence of prostate cancer after previous external beam radiotherapy. Between September 2008 and March 2013, 25 patients with local recurrence of previously irradiated prostate cancer were treated. The main eligibility criteria for salvage prostate HDR brachytherapy combined with interstitial hyperthermia were biopsy confirmed local recurrence and absence of nodal and distant metastases. All patients were treated with a dose of 30 Gy in 3 fractions at 21-day intervals. We performed 62 hyperthermia procedures out of 75 planned (83 %). The aim of the hyperthermia treatment was to heat the prostate to 41-43 C for 60 min. Toxicity for the organs of the genitourinary system and rectum was assessed according to the Common Terminology Criteria for Adverse Events (CTCAE, v. 4.03). Determination of subsequent biochemical failure was based on the Phoenix definition (nadir + 2 ng/ml). The median age was 71 years (range 62-83 years), the median initial PSA level was 16.3 ng/ml (range 6.37-64 ng/ml), and the median salvage PSA level was 2.8 ng/ml (1.044-25.346 ng/ml). The median follow-up was 13 months (range 4-48 months). The combination of HDR brachytherapy and IHT was well tolerated. The most frequent complications were nocturia, weak urine stream, urinary frequency, hematuria, and urgency. Grade 2 rectal hemorrhage was observed in 1 patient. No grade 3 or higher complications were observed. The 2-year Kaplan-Meier estimate of biochemical control after salvage treatment was 74 %. The PSA in 20 patients decreased below the presalvage level, while 11 patients achieved a PSA nadir < 0.5 ng/ml. All patients are still alive. Of the 7 patients who experienced biochemical failure, bone metastases were found in 2 patients. IHT in combination

  5. SU-C-202-02: A Comprehensive Evaluation of Adaptive Daily Planning for Cervical Cancer HDR Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Meerschaert, R; Paul, A; Zhuang, L [Department of Oncology, Radiation Oncology Division, Wayne State University School of Medicine, Detroit, MI (United States); Nalichowski, A [Department of Oncology, Radiation Oncology Division, Karmanos Cancer Institute, Detroit, MI (United States); Burmeister, J; Miller, A [Department of Oncology, Radiation Oncology Division, Wayne State University School of Medicine, Detroit, MI (United States); Department of Oncology, Radiation Oncology Division, Karmanos Cancer Institute, Detroit, MI (United States)

    2016-06-15

    Purpose: To evaluate adaptive daily planning for cervical cancer patients who underwent high-dose-rate intra-cavitary brachytherapy (HDR-ICBT). Methods: This study included 22 cervical cancer patients who underwent 5 fractions of HDR ICBT. Regions of interest (ROIs) including high-risk clinical tumor volume (HR-CTV) and organs-at-risk (OARs) were manually contoured on daily CT images. All patients were treated with adaptive daily plans, which involved ROI delineation and dose optimization at each treatment fraction. Single treatment plans were retrospectively generated by applying the first treatment fraction’s dwell times adjusted for decay and dwell positions of the applicator to subsequent treatment fractions. Various existing similarity metrics were calculated for the ROIs to quantify interfractional organ variations. A novel similarity score (JRARM) was established, which combined both volumetric overlap metrics (DSC, JSC, and RVD) and distance metrics (ASD, MSD, and RMSD). Linear regression was performed to determine a relationship between inter-fractional organ variations of various similarity metrics and D2cc variations from both plans. Wilcoxon Signed Rank Tests were used to assess adaptive daily plans and single plans by comparing EQD2 D2cc (α/β=3) for OARs. Results: For inter-fractional organ variations, the sigmoid demonstrated the greatest variations based on the JRARM and DSC similarity metrics. Comparisons between paired ROIs showed differences in JRARM scores and DSCs at each treatment fraction. RVD, MSD, and RMSD were found to be significantly correlated to D2cc variations for bladder and sigmoid. The comparison between plans found that adaptive daily planning provided lower EQD2 D2cc of OARs than single planning, specifically for the sigmoid (p=0.015). Conclusion: Substantial inter-fractional organ motion can occur during HDR-BT, which may significantly affect D2cc of OARs. Adaptive daily planning provides improved dose sparing for OARs

  6. SU-E-T-149: Brachytherapy Patient Specific Quality Assurance for a HDR Vaginal Cylinder Case

    Energy Technology Data Exchange (ETDEWEB)

    Barbiere, J; Napoli, J; Ndlovu, A [Hackensack Univ Medical Center, Hackensack, NJ (United States)

    2015-06-15

    Purpose: Commonly Ir-192 HDR treatment planning system commissioning is only based on a single absolute measurement of source activity supplemented by tabulated parameters for multiple factors without independent verification that the planned distribution corresponds to the actual delivered dose. The purpose on this work is to present a methodology using Gafchromic film with a statistically valid calibration curve that can be used to validate clinical HDR vaginal cylinder cases by comparing the calculated plan dose distribution in a plane with the corresponding measured planar dose. Methods: A vaginal cylinder plan was created with Oncentra treatment planning system. The 3D dose matrix was exported to a Varian Eclipse work station for convenient extraction of a 2D coronal dose plane corresponding to the film position. The plan was delivered with a sheet of Gafchromic EBT3 film positioned 1mm from the catheter using an Ir-192 Nucletron HDR source. The film was then digitized with an Epson 10000 XL color scanner. Film analysis is performed with MatLab imaging toolbox. A density to dose calibration curve was created using TG43 formalism for a single dwell position exposure at over 100 points for statistical accuracy. The plan and measured film dose planes were registered using a known dwell position relative to four film marks. The plan delivered 500 cGy to points 2 cm from the sources. Results: The distance to agreement of the 500 cGy isodose between the plan and film measurement laterally was 0.5 mm but can be as much as 1.5 mm superior and inferior. The difference between the computed plan dose and film measurement was calculated per pixel. The greatest errors up to 50 cGy are near the apex. Conclusion: The methodology presented will be useful to implement more comprehensive quality assurance to verify patient-specific dose distributions.

  7. Implementation of the technique of partial irradiation accelerated the breast with high doses (HDR) brachytherapy; Puesta en marcha de la tecnica de irradiacion parcial acelerada de la mama con braquterapia de alta tasa de dosis (HDR)

    Energy Technology Data Exchange (ETDEWEB)

    Molina Lopez, M. Y.; Pardo Perez, E.; Castro Novais, J.; Martinez Ortega, J.; Ruiz Maqueda, S.; Cerro Penalver, E. del

    2013-07-01

    The objective of this work is presents procedure carried out in our Centre for the implementation of the accelerated partial breast irradiation (APBI, accelerated partial-breast irradiation) with high-rate brachytherapy (HDR), using plastic tubes as applicators. Carried out measures, the evaluation of the dosimetric parameters analyzing and presenting the results. (Author)

  8. A phantom for verification of dwell position and time of a high dose rate brachytherapy source

    International Nuclear Information System (INIS)

    Madebo, M.; Kron, T.; Pillainayagam, J.; Franich, R.

    2012-01-01

    Accuracy of dwell position and reproducibility of dwell time are critical in high dose rate (HDR) brachytherapy. A phantom was designed to verify dwell position and dwell time reproducibility for an Ir-192 HDR stepping source using Computed Radiography (CR). The central part of the phantom, incorporating thin alternating strips of lead and acrylic, was used to measure dwell positions. The outer part of the phantom features recesses containing different absorber materials (lead, aluminium, acrylic and polystyrene foam), and was used for determining reproducibility of dwell times. Dwell position errors of <1 mm were easily detectable using the phantom. The effect of bending a transfer tube was studied with this phantom and no change of clinical significance was observed when varying the curvature of the transfer tube in typical clinical scenarios. Changes of dwell time as low as 0.1 s, the minimum dwell time of the treatment unit, could be detected by choosing dwell times over the four materials that produce identical exposure at the CR detector.

  9. Hydrogen dressings HDR

    International Nuclear Information System (INIS)

    Rosiak, J.M.

    1990-01-01

    Within the past several years new developments in biomaterials have enabled a significant progress in the healing of different kinds of wounds. One of such biomaterials is synthetic hydrogels. They can be formed by means of radiation technology which gives some advantages over chemical methods. Hydrogel dressings HDR have the shape of transparent foil, 3-4 mm thick, which contain over 90% water. They can be used for healing exuding wounds and especially burns, ulcers, bedsores and skin grafts. HDR dressings are sterile, transparent and mechanically resistant

  10. Relocation of a nucletron microselectron-HDR brachytherapy system

    Energy Technology Data Exchange (ETDEWEB)

    Bartrum, T; Tran, T; Freeman, N; Morales, J [St Vincents Hospital, Darlinghurst, NSW (Australia)

    2004-12-15

    Full text: For a period of four weeks, our clinical Nucletron microSelectron high dose rate (HDR) brachytherapy system was pulled out of clinical use and relocated to a new building. During this period decommission tests, de-wiring of the treatment unit and its associated safety system (such as radiation detector, emergency off circuits and door interlocks), transportation of all equipment, re-wiring of this equipment in the new location and recommission tests were carried out. The decommission and recommission test program was designed upon consultation with the manufacturer's (Nucletron) acceptance test procedures and work carried out by others. The ACPSEM tolerances for remote afterloaders was used as a guideline. In addition to mandatory dosimetry, positional, workstation database and safety tests, two Australian Standard compliance tests were carried out. The compliance tests involved one for remote afterloaders and another for treatment room design. This testing program was designed and implemented with the aim of ensuring ongoing safe delivery of brachytherapy doses to the patient. The testing program consisted of two parts. The first involved a series of decommissioning tests that consisted of dosimetry tests such as source and check cable positional accuracy and source calibration tests. In addition to these tests an inventory of standard plans, patient records and system configuration information was catalogued. The second part involved a series of recommission tests and involved carrying out dosimetry tests on the brachytherapy system (positional accuracy and calibration tests), simulating common treatment scenarios (prostate, cervical, vaginal and bile duct) and checking standard plans; patient records and system configuration had remained unchanged. During this period, other tests were carried out. These included Nucletron acceptance and preventative maintenance tests, Australian Standards compliance testing and integrity of network transfer of

  11. Relocation of a nucletron microselectron-HDR brachytherapy system

    International Nuclear Information System (INIS)

    Bartrum, T.; Tran, T.; Freeman, N.; Morales, J.

    2004-01-01

    Full text: For a period of four weeks, our clinical Nucletron microSelectron high dose rate (HDR) brachytherapy system was pulled out of clinical use and relocated to a new building. During this period decommission tests, de-wiring of the treatment unit and its associated safety system (such as radiation detector, emergency off circuits and door interlocks), transportation of all equipment, re-wiring of this equipment in the new location and recommission tests were carried out. The decommission and recommission test program was designed upon consultation with the manufacturer's (Nucletron) acceptance test procedures and work carried out by others. The ACPSEM tolerances for remote afterloaders was used as a guideline. In addition to mandatory dosimetry, positional, workstation database and safety tests, two Australian Standard compliance tests were carried out. The compliance tests involved one for remote afterloaders and another for treatment room design. This testing program was designed and implemented with the aim of ensuring ongoing safe delivery of brachytherapy doses to the patient. The testing program consisted of two parts. The first involved a series of decommissioning tests that consisted of dosimetry tests such as source and check cable positional accuracy and source calibration tests. In addition to these tests an inventory of standard plans, patient records and system configuration information was catalogued. The second part involved a series of recommission tests and involved carrying out dosimetry tests on the brachytherapy system (positional accuracy and calibration tests), simulating common treatment scenarios (prostate, cervical, vaginal and bile duct) and checking standard plans; patient records and system configuration had remained unchanged. During this period, other tests were carried out. These included Nucletron acceptance and preventative maintenance tests, Australian Standards compliance testing and integrity of network transfer of

  12. Radiobiological restrictions and tolerance doses of repeated single-fraction hdr-irradiation of intersecting small liver volumes for recurrent hepatic metastases

    Directory of Open Access Journals (Sweden)

    Wust Peter

    2010-05-01

    Full Text Available Abstract Background To assess radiobiological restrictions and tolerance doses as well as other toxic effects derived from repeated applications of single-fraction high dose rate irradiation of small liver volumes in clinical practice. Methods Twenty patients with liver metastases were treated repeatedly (2 - 4 times at identical or intersecting locations by CT-guided interstitial brachytherapy with varying time intervals. Magnetic resonance imaging using the hepatocyte selective contrast media Gd-BOPTA was performed before and after treatment to determine the volume of hepatocyte function loss (called pseudolesion, and the last acquired MRI data set was merged with the dose distributions of all administered brachytherapies. We calculated the BED (biologically equivalent dose for a single dose d = 2 Gy for different α/β values (2, 3, 10, 20, 100 based on the linear-quadratic model and estimated the tolerance dose for liver parenchyma D90 as the BED exposing 90% of the pseudolesion in MRI. Results The tolerance doses D90 after repeated brachytherapy sessions were found between 22 - 24 Gy and proved only slightly dependent on α/β in the clinically relevant range of α/β = 2 - 10 Gy. Variance analysis showed a significant dependency of D90 with respect to the intervals between the first irradiation and the MRI control (p 90 and the pseudolesion's volume. No symptoms of liver dysfunction or other toxic effects such as abscess formation occurred during the follow-up time, neither acute nor on the long-term. Conclusions Inactivation of liver parenchyma occurs at a BED of approx. 22 - 24 Gy corresponding to a single dose of ~10 Gy (α/β ~ 5 Gy. This tolerance dose is consistent with the large potential to treat oligotopic and/or recurrent liver metastases by CT-guided HDR brachytherapy without radiation-induced liver disease (RILD. Repeated small volume irradiation may be applied safely within the limits of this study.

  13. Exclusive radiation therapy of endometrial carcinoma using HDR Co-60 and Cf-252 sources

    International Nuclear Information System (INIS)

    Inciura, A.; Janulionis, E.; Atkocius, V.

    2000-01-01

    Although the main treatment of endometrial cancer is surgery, the gross spread of carcinoma in pelvis, elderly age of the patients and serious therapeutical diseases do not allow to operate 20% of the patients. For these patients only radical treatment decision is the combined radiation therapy. However, using various modalities of combined radiotherapy, the treatment results are not satisfying: only 52% of patients survive 5 years. There was implemented a new method of combined radiotherapy, using a three-channel applicator with two-positional locating of HDR Co-60 radioactive sources (group I). A new method of brachytherapy for endometrial carcinoma using a HDR Cf 252 sources (group II) was implemented too. For group I the medium total dose of the point A was 77.6 Gy, point I - 69.6 Gy. point 2 - 84.2 Gy, point B - 52.6 Gy, For group II the medium total dose the points A, 3 and 4 was 50 iGy. For group I 1 year overaII survival rate was 85%, 3 year - 73%. The 3 year survival rate for stage I was 79%, for stage II - 89%, for stage III - 33%. 3 year survival for highly virulent tumours - 56% and for low virulent (adenocarcinoma) - 82% was statistically different. The loss of tumour differentiation correlated with the worse prognosis: 3 year survival for G1 and G2 tumours was 92%, for G3 - 45%. Low hemoglobin level was also associated with low survival rate: 2 year survival for Hb≥120g/l was 85% and Hb<120g/l - 23%. 1 year survival rate was 87%, 3 year - 66%, 5 year - 58% for group II, treated with Cf-252. Radiation complications occurred for 13.8% of the patients for group I and 6.2% for group II. Good survival rates and a small number of complications sufficiently proofed treatment method. (author)

  14. SU-F-T-28: Evaluation of BEBIG HDR Co-60 After-Loading System for Skin Cancer Treatment Using Conical Surface Applicator

    Energy Technology Data Exchange (ETDEWEB)

    Safigholi, H; Soliman, A; Song, W Y [Department of Medical Physics, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, ON (Canada); Meigooni, A S [Department of Radiation Therapy, Comprehensive Cancer Center of Nevada, Las Vegas, NV (United States); Han, D [Departemt of Radiation Oncology, University of California San Francisco, San Francisco, CA (United States)

    2016-06-15

    Purpose: To evaluate the possibility of utilizing the BEBIG HDR 60Co remote after-loading system for malignant skin surface treatment using Monte Carlo (MC) simulation technique. Methods: First TG-43 parameters of BEBIG-Co-60 and Nucletron Ir-192-mHDR-V2 brachytherapy sources were simulated using MCNP6 code to benchmark the sources against the literature. Second a conical tungsten-alloy with 3-cm diameter of Planning-Target-Volume (PTV) at surface for use with a single stepping HDR source is designed. The HDR source is modeled parallel to treatment plane at the center of the conical applicator with a source surface distance (SSD) of 1.5-cm and a removable plastic end-cap with a 1-mm thickness. Third, MC calculated dose distributions from HDR Co-60 for conical surface applicator were compared with the simulated data using HDR Ir-192 source. The initial calculations were made with the same conical surface applicator (standard-applicator) dimensions as the ones used with the Ir-192 system. Fourth, the applicator wall-thickness for the Co-60 system was increased (doubled) to diminish leakage dose to levels received when using the Ir-192 system. With this geometry, percentage depth dose (PDD), and relative 2D-dose profiles in transverse/coronal planes were normalized at 3-mm prescription-depth evaluated along the central axis. Results: PDD for Ir-192 and Co-60 were similar with standard and thick-walled applicator. 2D-relative dose distribution of Co-60, inside the standard-conical-applicator, generated higher penumbra (7.6%). For thick-walled applicator, it created smaller penumbra (<4%) compared to Ir-192 source in the standard-conicalapplicator. Dose leakage outside of thick-walled applicator with Co-60 source was approximately equal (≤3%) with standard applicator using Ir-192 source. Conclusion: Skin cancer treatment with equal quality can be performed with Co-60 source and thick-walled conical applicators instead of Ir-192 with standard applicators. These conical

  15. Constraints in the use of repair half times and mathematical modelling for the clinical application of HDR and PDR treatment schedules as an alternative for LDR brachytherapy

    International Nuclear Information System (INIS)

    Pop, L.A.M.; Broek, J.F.C.M. van den; Visser, A.G.; Kogel, A.J. van der

    1996-01-01

    Using theoretical models based on radiobiological principles for the design of new treatment schedules for HDR and PDR brachytherapy, it is important to realise the impact of assumptions regarding the kinetics of repair. Extrapolations based on longer repair half times in a continuous LDR reference scheme may lead to the calculation of dangerously high doses for alternative HDR and PDR treatment schedules. We used the clinical experience obtained with conventional ERT and LDR brachytherapy in head and neck cancer as a clinical guideline to check the impact of the radiobiological parameters used. Biologically equivalent dose (BED) values for the in clinical practice of LDR brachytherapy recommended dose of 65-70 Gy (prescribed at a dose rate between 30-50 cGy/h) are calculated as a function of the repair half time. These BED values are compared with the biological effect of a clinical reference dose of conventional ERT with 2 Gy/day and complete repair between the fractions. From this comparison of LDR and ERT treatment schedules, a range of values for the repair half times of acute or late responding tissues is demarcated with a reasonable fit to the clinical data. For the acute effects (or tumor control) the best fits are obtained for repair half times of about 0.5 h, while for late effects the repair half times are at least 1 h and can be as high as 3 h. Within these ranges of repair half times for acute and late effects, the outcome of 'alternative' HDR or PDR treatment schedules are discussed. It is predominantly the late reacting normal tissue with the longer repair half time for which problems will be encountered and no or only marginal gain is to be expected of decreasing the dose rate per pulse in PDR brachytherapy

  16. Comparison of patient-reported acute urinary and sexual toxicity scores in a 6- versus 2-fraction course of high-dose-rate prostate brachytherapy monotherapy

    International Nuclear Information System (INIS)

    Ragab, Omar; Park, Sang-June; Zhang, Mingle; Wang, Jason; Velez, Maria; Demanes, David J.; Banerjee, Robyn; Patel, Shyamal; Kamrave, Mitchell

    2018-01-01

    To identify differences in acute urinary and sexual toxicity between a 6-fraction and 2-fraction high-dose-rate brachytherapy monotherapy regimen and correlate dosimetric constraints to short-term toxicity. A single institution retrospective study of 116 men with prostate cancer treated with HDR monotherapy from 2010 to 2015 was conducted. Eighty-one men had 7.25 Gy × 6-fractions and 35 men had 13.5 Gy × 2-fractions. Patients had two CT-planned implants spaced 1–2 weeks apart. Patient baseline characteristics, International Prostate Symptom Scores (IPSS) and Sexual Health Inventory for Men (SHIM) scores were collected pre-treatment and 3, 6 and 12 months post-implantation. Mixed effect modelling was undertaken to compare baseline, 1–6 month and 7–12 month scores between groups. Poisson regression analysis was performed to correlate dosimetric constraints with acute toxicity. There was no difference between baseline and post-implantation IPSS scores between 6-fraction and 2-fraction groups. SHIM scores for men treated with 6-fractions had a steeper decline at 1–6 months, but resolved at 7–12 months. Pre-treatment alpha-blocker use correlated with worse short-term acute urinary toxicity. Worsened SHIM score correlated with increasing age, diabetes mellitus and androgen-deprivation therapy. In a dosimetric analysis of outcomes, prostate V150 dose and bladder wall (D01.cc, D1cc, D2cc) dose correlated with increased IPSS score. No increased acute genitourinary or sexual dysfunction has been observed in men when transitioning from 6-fraction to 2-fraction HDR monotherapy. A dosimetric correlation was found between the V150 and bladder wall doses for acute urinary toxicity. Future research should continue to standardize and validate dose constraints for prostate HDR monotherapy patients.

  17. Application of Multiobjective Genetic Algorithms in Anatomy Based Dose Optimization in Brachytherapy and its Comparation with Deterministic Algorithms

    National Research Council Canada - National Science Library

    Milickovic, Natasa

    2001-01-01

    In High Dose Rate (HDR) brachytherapy the conventional dose optimization algorithms consider the multiple objectives in the form of an aggregate function which combines individual objectives into a single utility value...

  18. Toxicity and cosmetic result of partial breast high-dose-rate interstitial brachytherapy for conservatively operated early breast cancer

    International Nuclear Information System (INIS)

    Xiu Xia; Tripuraneni Prabhakar; Giap Huan; Lin Ray; Chu Colin

    2007-01-01

    Objective: Objective To study the method, side effects and cosmetic outcome of high- dose-rate (HDR) accelerated partial breast interstitial irradiation (APBI) alone in early stage breast cancer' after conservative surgery. Methods: From February 2002 to June 2003,47 breast cancer lesions from 46 patients suffering from stage I/II breast cancer were treated with HDR 192 Ir APBI after conservative surgery. All patients were over 40 year-old, with T1-2N0-1 (≤3 lymph nodes positive), surgical margin > 1-2 mm, but those having lobular or inflammatory breast cancer were excluded. HDR brachytherapy with 34 Gy, 10 fractions/5 days was used after surgery, toxic reaction and cosmetic outcome were observed in one month, 6 and 12 months respectively. Results: Follow up of 1846 months, 34 months was carried out for the whole group. During the treatment, acute reactions including: erythema, edema, tenderness and infection, all under I-II grade, none of III-IV grade were observed in 21 patients(46%); late toxicity reactions: skin fibrosis, breast tenderness, fat necrosis, and telangiectasia, totally 20 patients (43%) were observed: 2 patients in III grade but one patient received 6 cycle chemotherapy. The result of cosmetic outcome evaluation was excellent or good, at 6 months 95% and 12 months 98%, respectively, but there was no recurfence. Conclusions: Excellent and favorable cosmetic results are noted after APBI by interstitial alone. Acute and late reactions are few. Long term observation is necessary for the rate of' local control. (authors)

  19. SU-E-T-380: Evaluation of BEBIG HDR 60Co System for AccuBoost Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Zehtabian, M; Sina, S [Shiraz University, Shiraz, Fars (Iran, Islamic Republic of); Rivard, M [Tufts University School of Medicine, Boston, MA (United States); Meigooni, A Soleimani [Comprehensive Cancer Center of Nevada, Las Vegas, Nevada (United States)

    2015-06-15

    Purpose: In this project, the possibility of utilizing the BEBIG 60Co HDR system for AccuBoostTM treatment has been evaluated. Methods: Dose distributions in various breast sizes have been calculated for both Co-60 and Ir-192 sources using the MCNP5 code. These calculations were performed in breast tissues with thicknesses of 4cm, 6cm, and 8cm. The initial calculations were performed with the same applicator dimensions as the existing applicators used with the HDR Ir-192 system. The activity of the Co-60 source was selected such that the dose at the breast center was the same as the values from 192Ir. Then, the applicator thicknesses were increased to twice of those used with HDR Ir-192 system, for reducing skin and chest doses by Co-60 system. Dose to breast skin and chest wall were compared for both applicators types, with and without inclusion of a focusing cone at the applicator center. Results: The results showed that loading HDR Co-60 source inside the thin applicators impose higher doses to breast skin and chest wall compared to the 192Ir source. The area of the chest wall covered by 10Gy when treated by Co-60 with the thin and thick applicators, or treated by Ir-192 with thin applicator are 79cm2, 39cm2, and 3.8cm2, respectively. These values are reduced to 34cm2, 0cm2, and 0cm2 by using the focusing cone. It is worth noting that the breast skin areas covered by the 60Gy isodose line are 9.9cm2 and 7.8cm2 for Co-60 with the thin and thick applicators, respectively, while it is 20cm2 for Ir-192 when no focusing cone is present. These values are 0cm2, 0cm2, and 11cm2 in the presence of the focusing cone. Conclusion: The results indicate that using Co-60 with the thicker applicators is beneficial because of the higher half-life of Co-60, and the reduced maximum skin dose when compared with Ir-192.

  20. Perioperative fractionated high-dose rate brachytherapy for malignant bone and soft tissue tumors

    International Nuclear Information System (INIS)

    Koizumi, Masahiko; Inoue, Takehiro; Yamazaki, Hideya; Teshima, Teruki; Tanaka, Eiichi; Yoshida, Ken; Imai, Atsushi; Shiomi, Hiroya; Kagawa, Kazufumi; Araki, Nobuto; Kuratsu, Shigeyuki; Uchida, Atsumasa; Inoue, Toshihiko

    1999-01-01

    Purpose: To investigate the viability of perioperative fractionated HDR brachytherapy for malignant bone and soft tissue tumors, analyzing the influence of surgical margin. Methods and Materials: From July 1992 through May 1996, 16 lesions of 14 patients with malignant bone and soft tissue tumors (3 liposarcomas, 3 MFHs, 2 malignant schwannomas, 2 chordomas, 1 osteosarcoma, 1 leiomyosarcoma, 1 epithelioid sarcoma, and 1 synovial sarcoma) were treated at the Osaka University Hospital. The patients' ages ranged from 14 to 72 years (median: 39 years). Treatment sites were the pelvis in 6 lesions, the upper limbs in 5, the neck in 4, and a lower limb in 1. The resection margins were classified as intracapsular in 5 lesions, marginal in 5, and wide in 6. Postoperative fractionated HDR brachytherapy was started on the 4th-13th day after surgery (median: 6th day). The total dose was 40-50 Gy/7-10 fr/ 4-7 day (bid) at 5 or 10 mm from the source. Follow-up periods were between 19 and 46 months (median: 30 months). Results: Local control rates were 75% at 1 year and 48% in 2 years, and ultimate local control was achieved in 8 (50%) of 16 lesions. Of the 8 uncontrolled lesions, 5 (63%) had intracapsular (macroscopically positive) resection margins, and all the 8 controlled lesions (100%) had marginal (microscopically positive) or wide (negative) margins. Of the total, 3 patients died of both tumor and metastasis, 3 of metastasis alone, 1 of tumor alone, and 7 showed no evidence of disease. Peripheral nerve palsy was seen in one case after this procedure, but no infection or delayed wound healing caused by tubing or irradiation has occurred. Conclusion: Perioperative fractionated HDR brachytherapy is safe, well tolerated, and applicable to marginal or wide surgical margin cases

  1. Comparison of different treatment planning optimization methods for vaginal HDR brachytherapy with multichannel applicators: A reduction of the high doses to the vaginal mucosa is possible.

    Science.gov (United States)

    Carrara, Mauro; Cusumano, Davide; Giandini, Tommaso; Tenconi, Chiara; Mazzarella, Ester; Grisotto, Simone; Massari, Eleonora; Mazzeo, Davide; Cerrotta, Annamaria; Pappalardi, Brigida; Fallai, Carlo; Pignoli, Emanuele

    2017-12-01

    A direct planning approach with multi-channel vaginal cylinders (MVCs) used for HDR brachytherapy of vaginal cancers is particularly challenging. Purpose of this study was to compare the dosimetric performances of different forward and inverse methods used for the optimization of MVC-based vaginal treatments for endometrial cancer, with a particular attention to the definition of strategies useful to limit the high doses to the vaginal mucosa. Twelve postoperative vaginal HDR brachytherapy treatments performed with MVCs were considered. Plans were retrospectively optimized with three different methods: Dose Point Optimization followed by Graphical Optimization (DPO + GrO), Inverse Planning Simulated Annealing with two different class solutions as starting conditions (surflPSA and homogIPSA) and Hybrid Inverse Planning Optimization (HIPO). Several dosimetric parameters related to target coverage, hot spot extensions and sparing of organs at risk were analyzed to evaluate the quality of the achieved treatment plans. Dose homogeneity index (DHI), conformal index (COIN) and a further parameter quantifying the proportion of the central catheter loading with respect to the overall loading (i.e., the central catheter loading index: CCLI) were also quantified. The achieved PTV coverage parameters were highly correlated with each other but uncorrelated with the hot spot quantifiers. HomogIPSA and HIPO achieved higher DHIs and CCLIs and lower volumes of high doses than DPO + GrO and surflPSA. Within the investigated optimization methods, HIPO and homoglPSA showed the highest dose homogeneity to the target. In particular, homogIPSA resulted also the most effective in reducing hot spots to the vaginal mucosa. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  2. Traceable calibration of hospital 192Ir HDR sources

    International Nuclear Information System (INIS)

    Govinda Rajan, K.N.; Bhatt, B.C.; Pendse, A.M.; Kannan, V.

    2002-01-01

    Presently, no primary standard exists for the standardization of remote afterloading 192 Ir HDR sources. These sources are, therefore, being standardized by a few Secondary Standard Dosimetry Laboratories (SSDLs), in terms of Air Kerma Strength (AKS) or Reference Air Kerma Rate (RAKR) using a 0.6 cc Farmer type chamber, set up as an Interim Standard. These SSDLs offer calibration to well type of ionization chambers that are normally used by the hospitals for calibrating the 192 lr HDR source. Presently, in many countries, including India, well chambers are not commercially available. Nor do these countries offer any calibration service for 192 lr HDR source. With the result users make use of well chambers imported from different countries with their calibration traceable to the country of origin. Since no intercomparisons between these countries have been reported, the measurement consistency between hospitals becomes questionable. The problem is compounded by the fact that these chambers are used for several years without re-calibration since no calibration service is locally available. For instance, in India, the chambers have been in use in hospitals, since 1994, without a second calibration. Not all hospitals use the well chamber for the calibration of the 192 lr HDR source. Many hospitals make use of 0.6 cc chambers, in air, at short source to chamber distances, for measuring the AKS of the source. The latter method is prone to much larger inaccuracy due to the use of very short source to chamber distances without proper calibration jigs, use of 60 Co calibration factor for 192 Ir HDR source calibrations, neglecting correction factors for room scatter, fluence non-uniformity, use of arbitrary buildup factors for the buildup cap of the chamber etc. A comparison of the procedures used at hospitals revealed that various arbitrary methods are in use at hospitals. An indigenously developed well chamber was calibrated against a Reference Standard traceable to the

  3. Effects of prescription depth, cylinder size, treatment length, tip space, and curved end on doses in high-dose-rate vaginal brachytherapy

    International Nuclear Information System (INIS)

    Li Shidong; Aref, Ibrahim; Walker, Eleanor; Movsas, Benjamin

    2007-01-01

    Purpose: To determine the effects of the prescription depth, cylinder size, treatment length, tip space, and curved end on high-dose-rate vaginal brachytherapy (HDR-VBT) of endometrial cancer. Methods and Materials: Treatment plans were prescribed and optimized based on points at the cylinder surface or at 0.5-cm depth. Cylinder sizes ranging from 2 to 4 cm in diameter, and treatment lengths ranging from 3 to 8 cm were used. Dose points in various depths were precisely defined along the cylinder dome. The given dose and dose uniformity to a depth of interest were measured by the mean dose (MD) and standard deviation (SD), respectively, among the dose points belonging to the depth. Dose fall-off beyond the 0.5 cm treatment depth was determined by the ratio of MD at 0.75-cm depth to MD at 0.5-cm depth. Results: Dose distribution varies significantly with different prescriptions. The surface prescription provides more uniform doses at all depths in the target volume, whereas the 0.5-cm depth prescription creates larger dose variations at the cylinder surface. Dosimetric uncertainty increases significantly (>30%) with shorter tip space. Extreme hot (>150%) and cold spots (<60%) occur if no optimization points were placed at the curved end. Conclusions: Instead of prescribing to a depth of 0.5 cm, increasing the dose per fraction and prescribing to the surface with the exact surface points around the cylinder dome appears to be the optimal approach

  4. SU-F-BRF-09: A Non-Rigid Point Matching Method for Accurate Bladder Dose Summation in Cervical Cancer HDR Brachytherapy

    International Nuclear Information System (INIS)

    Chen, H; Zhen, X; Zhou, L; Zhong, Z; Pompos, A; Yan, H; Jiang, S; Gu, X

    2014-01-01

    Purpose: To propose and validate a deformable point matching scheme for surface deformation to facilitate accurate bladder dose summation for fractionated HDR cervical cancer treatment. Method: A deformable point matching scheme based on the thin plate spline robust point matching (TPSRPM) algorithm is proposed for bladder surface registration. The surface of bladders segmented from fractional CT images is extracted and discretized with triangular surface mesh. Deformation between the two bladder surfaces are obtained by matching the two meshes' vertices via the TPS-RPM algorithm, and the deformation vector fields (DVFs) characteristic of this deformation is estimated by B-spline approximation. Numerically, the algorithm is quantitatively compared with the Demons algorithm using five clinical cervical cancer cases by several metrics: vertex-to-vertex distance (VVD), Hausdorff distance (HD), percent error (PE), and conformity index (CI). Experimentally, the algorithm is validated on a balloon phantom with 12 surface fiducial markers. The balloon is inflated with different amount of water, and the displacement of fiducial markers is benchmarked as ground truth to study TPS-RPM calculated DVFs' accuracy. Results: In numerical evaluation, the mean VVD is 3.7(±2.0) mm after Demons, and 1.3(±0.9) mm after TPS-RPM. The mean HD is 14.4 mm after Demons, and 5.3mm after TPS-RPM. The mean PE is 101.7% after Demons and decreases to 18.7% after TPS-RPM. The mean CI is 0.63 after Demons, and increases to 0.90 after TPS-RPM. In the phantom study, the mean Euclidean distance of the fiducials is 7.4±3.0mm and 4.2±1.8mm after Demons and TPS-RPM, respectively. Conclusions: The bladder wall deformation is more accurate using the feature-based TPS-RPM algorithm than the intensity-based Demons algorithm, indicating that TPS-RPM has the potential for accurate bladder dose deformation and dose summation for multi-fractional cervical HDR brachytherapy. This work is supported

  5. NPIP: A skew line needle configuration optimization system for HDR brachytherapy

    International Nuclear Information System (INIS)

    Siauw, Timmy; Cunha, Adam; Berenson, Dmitry; Atamtürk, Alper; Hsu, I-Chow; Goldberg, Ken; Pouliot, Jean

    2012-01-01

    Purpose: In this study, the authors introduce skew line needle configurations for high dose rate (HDR) brachytherapy and needle planning by integer program (NPIP), a computational method for generating these configurations. NPIP generates needle configurations that are specific to the anatomy of the patient, avoid critical structures near the penile bulb and other healthy structures, and avoid needle collisions inside the body. Methods: NPIP consisted of three major components: a method for generating a set of candidate needles, a needle selection component that chose a candidate needle subset to be inserted, and a dose planner for verifying that the final needle configuration could meet dose objectives. NPIP was used to compute needle configurations for prostate cancer data sets from patients previously treated at our clinic. NPIP took two user-parameters: a number of candidate needles, and needle coverage radius, δ. The candidate needle set consisted of 5000 needles, and a range of δ values was used to compute different needle configurations for each patient. Dose plans were computed for each needle configuration. The number of needles generated and dosimetry were analyzed and compared to the physician implant. Results: NPIP computed at least one needle configuration for every patient that met dose objectives, avoided healthy structures and needle collisions, and used as many or fewer needles than standard practice. These needle configurations corresponded to a narrow range of δ values, which could be used as default values if this system is used in practice. The average end-to-end runtime for this implementation of NPIP was 286 s, but there was a wide variation from case to case. Conclusions: The authors have shown that NPIP can automatically generate skew line needle configurations with the aforementioned properties, and that given the correct input parameters, NPIP can generate needle configurations which meet dose objectives and use as many or fewer

  6. SU-E-T-263: Point Dose Variation Using a Single Ir-192 HDR Brachytherapy Plan for Two Treatments with a Single Tandem-Ovoid Insertion for Cervical Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Liang, X; Morrill, S; Hardee, M; Han, E; Penagaricano, J; Zhang, X; Vaneerat, R [University of Arkansas Medical Science, Little Rock, AR (United States)

    2014-06-01

    Purpose: To evaluate the point dose variations between Ir-192 HDR treatments on two consecutive days using a single tandem-ovoid insertion without replanning in cervical cancer patients. Methods: This study includes eleven cervical cancer patients undergoing HDR brachytherapy with a prescribed dose of 28 Gy in 4 fractions. Each patient had two tandemovoid insertions one week apart. Each insertion was treated on consecutive days with rescanning and replanning prior to each treatment. To study the effect of no replanning for day 2 treatments, the day 1 plan dwell position and dwell time with decay were applied to the day 2 CT dataset. The point dose variations on the prescription point H (defined according to American Brachytherapy Society), and normal tissue doses at point B, bladder, rectum and vaginal mucosa (based on ICRU Report 38) were obtained. Results: Without replanning, the mean point H dose variation was 4.6 ± 10.7% on the left; 2.3 ± 2.9% on the right. The mean B point variation was 3.8 ± 4.9% on the left; 3.6 ± 4.7% on the right. The variation in the left vaginal mucosal point was 12.2 ± 10.7%; 9.5 ± 12.5% on the right; the bladder point 5.5 ± 7.4%; and the rectal point 7.9 ± 9.1%. Conclusion: Without replanning, there are variations both in the prescription point and the normal tissue point doses. The latter can vary as much as 10% or more. This is likely due to the steep dose gradient from brachytherapy compounded by shifts in the positions of the applicator in relationship to the patients anatomy. Imaging prior to each treatment and replanning ensure effective and safe brachytherapy are recommended.

  7. Endoluminal high dose rate brachytherapy in the treatment of primary and recurrent bronchogenic tree malignancies

    Directory of Open Access Journals (Sweden)

    Maria Fortunato

    2009-03-01

    Full Text Available Introduction: Locally advanced tumours as the initial form of presentation of tumours in the bronchial tree are not a rare event. Bronchogenic recurrence is frequent in the natural history of some tumours. The choice of therapeutic options from the raft available depends on such variables as initial therapy, place of recurrence, symptoms and patient's physical status. Aim: To demonstrate the advantages of endoluminal brachytherapy (EBT with high dose rate (HDR in primary and recurrent tumour of the bronchial tree. Material and methods: A retrospective study of seven patients (pts with primary tumours of the colon, trachea and lung. Tracheobronchial recurrence (trachea, two pts, bronchus, five pts occurred betweenMarch 2003 and September 2004. Patients under-went EBT with HDR for primary or recurrent therapy in association with external radiotherapy, laser therapy and chemotherapy with palliative or curative intention. EBT with HDR doses of 5 to 7 Gy in 2 to 4 fractions at 1 cm from the source axis were given. Treatment included endoluminal application of Ir192 with a French 6 catheter. Results: There was symptomatic relief related to reduction of tumour in six of the seven patients treated. In one of the six patients studied, there was progression of the local disease between the second and third fractions of the treatment (obstruction of the trachea. In a mean follow up of 17 (2-40 months between EBT and this study, three patients are alive, one has no evidence of disease while two have had bronchial recurrence, four patients have died, one after massive haemoptysis and three due to disease progression. Discussion and conclusions: Patients undergoing brachytherapy for symptomatic primary tumours or endobronchial recurrence show good tolerance, important symptom relief and improved quality of life. Despite the small size of our sample, it is clear that EBT with HDR plays an important role in the palliative/curative treatment of these patients

  8. Effect of rectal enemas on rectal dosimetric parameters during high-dose-rate vaginal cuff brachytherapy. A prospective trial

    Energy Technology Data Exchange (ETDEWEB)

    Sabater, Sebastia; Andres, Ignacio; Sevillano, Marimar; Berenguer, Roberto; Aguayo, Manuel; Villas, Maria Victoria [Complejo Hospitalario Universitario de Albacete (CHUA), Department of Radiation Oncology, Albacete (Spain); Gascon, Marina; Arenas, Meritxell [Hospital Universitari Sant Joan, Department of Radiation Oncology, Reus (Spain); Rovirosa, Angeles; Camacho-Lopez, Cristina [University of Barcelona, IDIBAPS, Gynecological Cancer Unit, Radiation Oncology Department, ICMHO, Hospital Clinic, Barcelona (Spain)

    2016-04-15

    To evaluate the effects of rectal enemas on rectal doses during postoperative high-dose-rate (HDR) vaginal cuff brachytherapy (VCB). This prospective trial included 59 patients. Two rectal cleansing enemas were self-administered before the second fraction, and fraction 1 was considered the basal status. Dose-volume histogram (DVH) values were generated for the rectum and correlated with rectal volume variation. Statistical analyses used paired and unpaired t-tests. Despite a significant 15 % reduction in mean rectal volume (44.07 vs. 52.15 cc, p = 0.0018), 35.6 % of patients had larger rectums after rectal enemas. No significant rectal enema-related DVH differences were observed compared to the basal data. Although not statistically significant, rectal cleansing-associated increases in mean rectal DVH values were observed: D{sub 0.1} {sub cc}: 6.6 vs. 7.21 Gy; D{sub 1} {sub cc}: 5.35 vs. 5.52 Gy; D{sub 2} {sub cc}: 4.67 vs. 4.72 Gy, before and after rectal cleaning, respectively (where D{sub x} {sub cc} is the dose to the most exposed x cm {sup 3}). No differences were observed in DVH parameters according to rectal volume increase or decrease after the enema. Patients whose rectal volume increased also had significantly larger DVH parameters, except for D{sub 5} {sub %}, D{sub 25} {sub %}, and D{sub 50} {sub %}. In contrast, in patients whose rectal volume decreased, significance was only seen for D{sub 25} {sub %} and D{sub 50} {sub %} (D{sub x} {sub %} dose covering x % of the volume). In the latter patients, nonsignificant reductions in D{sub 2} {sub cc}, D{sub 5} {sub cc} and V{sub 5} {sub Gy} (volume receiving at least 5 Gy) were observed. The current rectal enemas protocol was ineffective in significantly modifying rectal DVH parameters for HDR-VCB. (orig.) [German] Beurteilung der Auswirkungen von rektalen Dosen waehrend postoperativer High-Dose-Rate-(HDR-)Brachytherapie an der Scheidenmanschette (''vaginal cuff brachytherapy'', VCB). An

  9. High-dose-rate brachytherapy with two or three fractions as monotherapy in the treatment of locally advanced prostate cancer

    International Nuclear Information System (INIS)

    Hoskin, Peter; Rojas, Ana; Ostler, Peter; Hughes, Robert; Alonzi, Roberto; Lowe, Gerry; Bryant, Linda

    2014-01-01

    Background: To evaluate late urinary (GU) and gastrointestinal (GI) adverse events (AEs) and biochemical control of disease after high-dose rate brachytherapy (HDR-BT) in locally advanced prostate cancer. Patients and methods: 227 consecutive patients were treated with 3 × 10.5 Gy (n = 109) or 2 × 13 Gy (n = 118) HDR-BT alone. Biochemical failure was assessed using the Phoenix definition of PSA nadir + 2 μg/l and late AEs using the RTOG scoring system and the International Prostate Symptom Score (IPSS). Results: Kaplan–Meier estimates and prevalence of late events indicate that urinary, bowel and IPSS symptoms are higher after 31.5 Gy than after 26 Gy, however differences are significant only for Grade 1 and 2 urinary toxicity. Kaplan–Meier estimates of morbidity are consistently and considerably higher than time-point estimates of prevalence; which reflects the transient nature of most symptoms. At 3 years 93% and 97% of patients treated with 26 and 31.5 Gy, respectively, were free from biochemical relapse (p = 0.5) and 91% for the latter regimen at 5 years. In univariate and multivariate analysis risk-category was the only significant predictor of relapse (p < 0.03). Conclusion: These HDR-BT schedules achieved high levels of biochemical control of disease in patients with advanced prostate cancer with few severe complications seen throughout the first 3 years

  10. A comprehensive study on HDR brachytherapy treatments of cervical cancers: using the first Co-60 BEBIG Multisource Unit in Bangladesh

    Directory of Open Access Journals (Sweden)

    Naheed Rukhsana

    2011-07-01

    Full Text Available Purpose: The report presents an extraordinary synthesis of customer acceptance procedures (CAP, quality assurance tests (QA in the treatment of cervical cancer patients, using the first Co-60 Multisource Unit® in Bangladesh. The QA and commissioning required measurements and emergency tests verifying the functional limits of parameters acceptable for the new HDR afterloader. Acceptable limits were: 1 the deviation between specified and measured source strength: ± 3%; 2 the positional accuracy and uniformity: ± 1 mm; 3 the temporal accuracy (i.e. timer error and linearity and end error: ± 1% or 30 sec.; 4 treatment planning system (digitizer and localization software: ± 3% or 1 mm; 5 the distance from line to first dwell position and all the others: 5 mm and 10 mm (± 1 mm. Material and methods: Till February 2011, 47 patients were treated with HDR with more than 140 insertions applied. Amongst them, 12 patients were in stage IIB and IIIB, 22 were postoperative (IA and IB while the remaining 13 patients were with unknown stage. All the cases with stage IIB and IIIB received concurrent chemo-radiation and brachytherapy. Postoperative patients received EBRT (50 Gy and HDR according to the institutional protocol. CT scans were completed before HDR-plus planning with a good reproducibility (± 2% and were documented in repeating the plan for the same set up of a patient. Absorbed dose (Gy to a point P, at a distance of “r†in centimeters from a source of the Reference Air Kerma Rate (RAKR has been utilized for the QA of the source, where source strength measurement was accomplished. Results: All methods and analysis applicable to the QA and commissioning of Co-60 have been investigated and systematically analyzed, measured and documented before the treatment of a patient. Studies and safety requirements of this HDR remote afterloader were carried out. Acceptance and the QA were imperative to justify functionality and dependability in

  11. A comprehensive study on HDR brachytherapy treatments of cervical cancers: using the first Co-60 BEBIG Multisource Unit in Bangladesh.

    Science.gov (United States)

    Malik, Sadiq R; Banu, Parvin A; Rukhsana, Naheed

    2011-06-01

    The report presents an extraordinary synthesis of customer acceptance procedures (CAP), quality assurance tests (QA) in the treatment of cervical cancer patients, using the first Co-60 Multisource Unit ® in Bangladesh. The QA and commissioning required measurements and emergency tests verifying the functional limits of parameters acceptable for the new HDR afterloader. Acceptable limits were: 1) the deviation between specified and measured source strength: ± 3%; 2) the positional accuracy and uniformity: ± 1 mm; 3) the temporal accuracy (i.e. timer error and linearity and end error): ± 1% or 30 sec.; 4) treatment planning system (digitizer and localization software): ± 3% or 1 mm; 5) the distance from line to first dwell position and all the others: 5 mm and 10 mm (± 1 mm). Till February 2011, 47 patients were treated with HDR with more than 140 insertions applied. Amongst them, 12 patients were in stage IIB and IIIB, 22 were postoperative (IA and IB) while the remaining 13 patients were with unknown stage. All the cases with stage IIB and IIIB received concurrent chemo-radiation and brachytherapy. Postoperative patients received EBRT (50 Gy and HDR) according to the institutional protocol. CT scans were completed before HDR-plus planning with a good reproducibility (± 2%) and were documented in repeating the plan for the same set up of a patient. Absorbed dose (Gy) to a point P, at a distance of "r" in centimeters from a source of the Reference Air Kerma Rate (RAKR) has been utilized for the QA of the source, where source strength measurement was accomplished. All methods and analysis applicable to the QA and commissioning of Co-60 have been investigated and systematically analyzed, measured and documented before the treatment of a patient. Studies and safety requirements of this HDR remote afterloader were carried out. Acceptance and the QA were imperative to justify functionality and dependability in delivering the treatment. Implications of these studies

  12. Computed Tomography–Guided Interstitial High-Dose-Rate Brachytherapy in Combination With Regional Positive Lymph Node Intensity-Modulated Radiation Therapy in Locally Advanced Peripheral Non–Small Cell Lung Cancer: A Phase 1 Clinical Trial

    Energy Technology Data Exchange (ETDEWEB)

    Xiang, Li; Zhang, Jian-wen; Lin, Sheng; Luo, Hui-Qun; Wen, Qing-Lian; He, Li-Jia; Shang, Chang-Ling; Ren, Pei-Rong; Yang, Hong-Ru; Pang, Hao-Wen; Yang, Bo; He, Huai-Lin [Department of Oncology, Affiliated Hospital of Luzhou Medical College, Luzhou (China); Chen, Yue, E-mail: chenyue5523@126.com [Department of Nuclear Medicine, Affiliated Hospital of Luzhou Medical College, Luzhou (China); Wu, Jing-Bo, E-mail: wjb6147@163.com [Department of Oncology, Affiliated Hospital of Luzhou Medical College, Luzhou (China)

    2015-08-01

    Purpose: To assess the technical safety, adverse events, and efficacy of computed tomography (CT)-guided interstitial high-dose-rate (HDR) brachytherapy in combination with regional positive lymph node intensity modulated radiation therapy in patients with locally advanced peripheral non–small cell lung cancer (NSCLC). Methods and Materials: Twenty-six patients with histologically confirmed NSCLC were enrolled in a prospective, officially approved phase 1 trial. Primary tumors were treated with HDR brachytherapy. A single 30-Gy dose was delivered to the 90% isodose line of the gross lung tumor volume. A total dose of at least 70 Gy was administered to the 95% isodose line of the planning target volume of malignant lymph nodes using 6-MV X-rays. The patients received concurrent or sequential chemotherapy. We assessed treatment efficacy, adverse events, and radiation toxicity. Results: The median follow-up time was 28 months (range, 7-44 months). There were 3 cases of mild pneumothorax but no cases of hemothorax, dyspnea, or pyothorax after the procedure. Grade 3 or 4 acute hematologic toxicity was observed in 5 patients. During follow-up, mild fibrosis around the puncture point was observed on the CT scans of 2 patients, but both patients were asymptomatic. The overall response rates (complete and partial) for the primary mass and positive lymph nodes were 100% and 92.3%, respectively. The 1-year and 2-year overall survival (OS) rates were 90.9% and 67%, respectively, with a median OS of 22.5 months. Conclusion: Our findings suggest that HDR brachytherapy is safe and feasible for peripheral locally advanced NSCLC, justifying a phase 2 clinical trial.

  13. Accuracy of applicator tip reconstruction in MRI-guided interstitial 192Ir-high-dose-rate brachytherapy of liver tumors

    International Nuclear Information System (INIS)

    Wybranski, Christian; Eberhardt, Benjamin; Fischbach, Katharina; Fischbach, Frank; Walke, Mathias; Hass, Peter; Röhl, Friedrich-Wilhelm; Kosiek, Ortrud; Kaiser, Mandy; Pech, Maciej; Lüdemann, Lutz; Ricke, Jens

    2015-01-01

    Background and purpose: To evaluate the reconstruction accuracy of brachytherapy (BT) applicators tips in vitro and in vivo in MRI-guided 192 Ir-high-dose-rate (HDR)-BT of inoperable liver tumors. Materials and methods: Reconstruction accuracy of plastic BT applicators, visualized by nitinol inserts, was assessed in MRI phantom measurements and in MRI 192 Ir-HDR-BT treatment planning datasets of 45 patients employing CT co-registration and vector decomposition. Conspicuity, short-term dislocation, and reconstruction errors were assessed in the clinical data. The clinical effect of applicator reconstruction accuracy was determined in follow-up MRI data. Results: Applicator reconstruction accuracy was 1.6 ± 0.5 mm in the phantom measurements. In the clinical MRI datasets applicator conspicuity was rated good/optimal in ⩾72% of cases. 16/129 applicators showed not time dependent deviation in between MRI/CT acquisition (p > 0.1). Reconstruction accuracy was 5.5 ± 2.8 mm, and the average image co-registration error was 3.1 ± 0.9 mm. Vector decomposition revealed no preferred direction of reconstruction errors. In the follow-up data deviation of planned dose distribution and irradiation effect was 6.9 ± 3.3 mm matching the mean co-registration error (6.5 ± 2.5 mm; p > 0.1). Conclusion: Applicator reconstruction accuracy in vitro conforms to AAPM TG 56 standard. Nitinol-inserts are feasible for applicator visualization and yield good conspicuity in MRI treatment planning data. No preferred direction of reconstruction errors were found in vivo

  14. An analysis of acute complications and perioperative morbidity from high dose rate brachytherapy in the treatment of gynecological malignancies

    Energy Technology Data Exchange (ETDEWEB)

    Sarkaria, Jann N; Petereit, Daniel G; Kinsella, Timothy J; Buchler, Dolores A

    1995-07-01

    Purpose: To evaluate the acute morbidity and mortality for high dose-rate (HDR) brachytherapy performed in an outpatient setting in the treatment of gynecological malignancies, and to identify possible risk factors for adverse outcomes. Materials and Methods: One hundred seventy-one patients with cervical (n=129) or uterine (n=42) carcinoma with an intact uterus were evaluated and treated from August 1989 through December 1994, with at least part of their therapy delivered with intracavitary HDR {sup 192}Ir radiation. A total of 830 ICR insertions were performed with greater than 95% done on an outpatient basis under heavy intravenous sedation using fentanyl and midazolam. Perioperative morbidity and mortality were recorded for any event occurring within 30 days of the completion of therapy. Anesthesia risk was evaluated retrospectively in all patients based on the American Society of Anesthesiologists' (ASA) Physical Class System. Results: The uterine patients, many treated with radiation alone because of morbid obesity or medical inoperability, had a significantly higher perioperative morbidity and mortality rate as compared to the cervix patient cohort. Thirteen of the 42 (31%) uterine patients and 8 of the 129 (6%) cervix patients required hospitalization within 30 days of treatment completion (p<0.001). Four of these uterine patients (10%) died of medical complications related to therapy or disease progression within 30 days, while 2 of the cervix patients (1.5%) died: one from disease progression and the other due to radiation enteritis (p=0.05). Preliminary analysis suggests that uterine patients with a high anesthesia risk score (ASA>2) experienced greater morbidity and mortality, while the best predictor of complications in the cervix patients was age greater than 70 years. For the entire cohort of patients, no correlation was found between the 30 day morbidity and mortality and the doses of fentanyl and midazolam used or the length of the procedure

  15. 3D dosimetry in HDR brachytherapy resonance imaging nuclear magnetic (b= 0.2 t) using a base acrylic gel (MAGIC)

    International Nuclear Information System (INIS)

    Batista Hernandez, Guillermo; Velez, Graciela R.; Schurrer, Clemar

    2009-01-01

    Dosimetry gels using magnetic resonance imaging (MRI) has been extended in recent literature. Our study presents the preparation, calibration IRM of acrylic gel (MAGIC) and its application in measuring dose in a 3D distribution HDR Brachytherapy with 192Ir source. The first gels used were the type Fricke gels based on the relationship of dose and time T1 relaxation. In 2001, Fong presented a new normoxic gel known as MAGIC whose main components are Methacrylic Acid (polymerizing), and Hydroquinone (inhibitor of self-curing) based on the relationship of dose and T2 relaxation time. Subsequent studies make changes in the concentrations component of the MAGIC (Methacrylic Acid and Hydroquinone in particular) to observe the behavior of the sensitivity of the gel with respect to its components and beam magnetic resonance equipment using magnetic fields higher to 0.5 T. This is done with equipment available to the staff of a Radiotherapy clinic setting. MAGIC gel is prepared according to composition by Crescenti (6% methacrylic acid), is calibrated with a 60Co unit TERADI INVAP 8002c (Argentina). Was raised shooting in a Siemens MRI scanner of 0.2 T Magnetom Concerto irradiated with a team of Brachytherapy High Dose Rate (HDR) Micro selectron Nucletron's V2 HDR for comparison with dose distributions provided by the planning system from Nucletron PLATO Sunrise. Was obtained a calibration curve for doses ranging from 0 to 8.0 Gy and a field strength 0.2 T magnetic We compared the sensitivity obtained in our calibration (Slope of the calibration curve) with those presented in the literature. Two phantoms were prepared for measurement in brachytherapy: a PMMA and a PVC. It was noted that MAGIC gel reacts chemically with PMMA and cured prior to irradiation. The phantom of PVC (no reactions) were irradiated with Micro selectron equipment and measured the dose distribution in 3D MRI. Were measured doses at the points specified by the Planning System and PLATO Sunrise compared

  16. Preoperative radiotherapy with high dose rate brachytherapy in the treatment of stage IIB cervix cancer. A retrospective analysis of histological specimens

    International Nuclear Information System (INIS)

    Ferrigno, Robson; Trippe, N; Novaes, P.E.; Brandani, I.B.; Hanriot, R.; Souza, L.M.; Pellizzon, A.C.; Salvajoli, J.V.; Baraldi, H.E.; Maia, M.A.; Fogaroli, R.C.

    1996-01-01

    Purpose/Objective: To evaluate the histological specimens of the stage IIB cervix cancer patients who were treated by preoperative radiotherapy with external beam radiotherapy (EBRT) and high dose rate (HDR) brachytherapy. Materials and Methods: From August 1992 to August 1995, 32 patients with stage IIB cervix cancer were underwent to preoperative radiotherapy. All patients received EBRT at the whole pelvis with total dose of 45Gy in 25 fractions of 1,8Gy through a 4 MV linear accelerator. The HDR brachytherapy was realized through a Micro-Selectron device, working with Iridium-192 with initial activity of 10 Ci. The prescribed dose was 6,0Gy at point A, defined by the Manchester, system in 2 weekly insertions during the course of EBRT. The insertions were done by the Fletcher colpostats in association with intrauterine tandem. Four to six weeks after the end of radiotherapy, the patients were underwent to Total Hysterectomy and Salpingoforectomy through Piver second level technique. The uterine specimens were histologically analysed with attention to residual disease at the cervix and lymph nodes status. Results: The histological analysis showed that 19 (59,4%) patients had no residual tumor at the cervix while 13 (40,6%) had microscopic residual tumor. The lymph nodes were negative in 30 (93,8%) patients and positive in 2 (6,3%). All positive lymph nodes patients also had microscopic residual tumor at the cervix. With the follow up ranging from six to 42 months and medium of 21 months, 29 (90,6%) patients are alive with no evidence of disease, one (5,6%) is alive with local recurrence and two (6,2%) have died due to the progression of local disease. Of the 19 patients with negative specimens, 18 (94,7%) are alive with no evidence of disease and of the 13 patients with positive specimens, 11 (84,6%) are alive with no evidence of disease. Local recurrence occurred in two patients with positive specimens and in one with negative. These differences are not

  17. Effect of rectal enemas on rectal dosimetric parameters during high-dose-rate vaginal cuff brachytherapy. A prospective trial

    International Nuclear Information System (INIS)

    Sabater, Sebastia; Andres, Ignacio; Sevillano, Marimar; Berenguer, Roberto; Aguayo, Manuel; Villas, Maria Victoria; Gascon, Marina; Arenas, Meritxell; Rovirosa, Angeles; Camacho-Lopez, Cristina

    2016-01-01

    To evaluate the effects of rectal enemas on rectal doses during postoperative high-dose-rate (HDR) vaginal cuff brachytherapy (VCB). This prospective trial included 59 patients. Two rectal cleansing enemas were self-administered before the second fraction, and fraction 1 was considered the basal status. Dose-volume histogram (DVH) values were generated for the rectum and correlated with rectal volume variation. Statistical analyses used paired and unpaired t-tests. Despite a significant 15 % reduction in mean rectal volume (44.07 vs. 52.15 cc, p = 0.0018), 35.6 % of patients had larger rectums after rectal enemas. No significant rectal enema-related DVH differences were observed compared to the basal data. Although not statistically significant, rectal cleansing-associated increases in mean rectal DVH values were observed: D 0.1 cc : 6.6 vs. 7.21 Gy; D 1 cc : 5.35 vs. 5.52 Gy; D 2 cc : 4.67 vs. 4.72 Gy, before and after rectal cleaning, respectively (where D x cc is the dose to the most exposed x cm 3 ). No differences were observed in DVH parameters according to rectal volume increase or decrease after the enema. Patients whose rectal volume increased also had significantly larger DVH parameters, except for D 5 % , D 25 % , and D 50 % . In contrast, in patients whose rectal volume decreased, significance was only seen for D 25 % and D 50 % (D x % dose covering x % of the volume). In the latter patients, nonsignificant reductions in D 2 cc , D 5 cc and V 5 Gy (volume receiving at least 5 Gy) were observed. The current rectal enemas protocol was ineffective in significantly modifying rectal DVH parameters for HDR-VCB. (orig.) [de

  18. TU-F-BRF-02: MR-US Prostate Registration Using Patient-Specific Tissue Elasticity Property Prior for MR-Targeted, TRUS-Guided HDR Brachytherapy

    International Nuclear Information System (INIS)

    Yang, X; Rossi, P; Ogunleye, T; Jani, A; Curran, W; Liu, T

    2014-01-01

    Purpose: High-dose-rate (HDR) brachytherapy has become a popular treatment modality for prostate cancer. Conventional transrectal ultrasound (TRUS)-guided prostate HDR brachytherapy could benefit significantly from MR-targeted, TRUS-guided procedure where the tumor locations, acquired from the multiparametric MRI, are incorporated into the treatment planning. In order to enable this integration, we have developed a MR-TRUS registration with a patient-specific biomechanical elasticity prior. Methods: The proposed method used a biomechanical elasticity prior to guide the prostate volumetric B-spline deformation in the MRI and TRUS registration. The patient-specific biomechanical elasticity prior was generated using ultrasound elastography, where two 3D TRUS prostate images were acquired under different probe-induced pressures during the HDR procedure, which takes 2-4 minutes. These two 3D TRUS images were used to calculate the local displacement (elasticity map) of two prostate volumes. The B-spline transformation was calculated by minimizing the Euclidean distance between the normalized attribute vectors of the prostate surface landmarks on the MR and TRUS. This technique was evaluated through two studies: a prostate-phantom study and a pilot study with 5 patients undergoing prostate HDR treatment. The accuracy of our approach was assessed through the locations of several landmarks in the post-registration and TRUS images; our registration results were compared with the surface-based method. Results: For the phantom study, the mean landmark displacement of the proposed method was 1.29±0.11 mm. For the 5 patients, the mean landmark displacement of the surface-based method was 3.25±0.51 mm; our method, 1.71±0.25 mm. Therefore, our proposed method of prostate registration outperformed the surfaced-based registration significantly. Conclusion: We have developed a novel MR-TRUS prostate registration approach based on patient-specific biomechanical elasticity prior

  19. Novel high dose rate lip brachytherapy technique to improve dose homogeneity and reduce toxicity by customized mold

    International Nuclear Information System (INIS)

    Feldman, Jon; Appelbaum, Limor; Sela, Mordechay; Voskoboinik, Ninel; Kadouri, Sarit; Weinberger, Jeffrey; Orion, Itzhak; Meirovitz, Amichay

    2014-01-01

    The purpose of this study is to describe a novel brachytherapy technique for lip Squamous Cell Carcinoma, utilizing a customized mold with embedded brachytherapy sleeves, which separates the lip from the mandible, and improves dose homogeneity. Seven patients with T2 lip cancer treated with a “sandwich†technique of High Dose Rate (HDR) brachytherapy to the lip, consisting of interstitial catheters and a customized mold with embedded catheters, were reviewed for dosimetry and outcome using 3D planning. Dosimetric comparison was made between the “sandwich†technique to “classic†– interstitial catheters only plan. We compared dose volume histograms for Clinical Tumor Volume (CTV), normal tissue “hot spots†and mandible dose. We are reporting according to the ICRU 58 and calculated the Conformal Index (COIN) to show the advantage of our technique. The seven patients (ages 36–81 years, male) had median follow-up of 47 months. Four patients received Brachytherapy and External Beam Radiation Therapy, 3 patients received brachytherapy alone. All achieved local control, with excellent esthetic and functional results. All patients are disease free. The Customized Mold Sandwich technique (CMS) reduced the high dose region receiving 150% (V150) by an average of 20% (range 1–47%), The low dose region (les then 90% of the prescribed dose) improved by 73% in average by using the CMS technique. The COIN value for the CMS was in average 0.92 as opposed to 0.88 for the interstitial catheter only. All differences (excluding the low dose region) were statistically significant. The CMS technique significantly reduces the high dose volume and increases treatment homogeneity. This may reduce the potential toxicity to the lip and adjacent mandible, and results in excellent tumor control, cosmetic and functionality

  20. FZUImageReg: A toolbox for medical image registration and dose fusion in cervical cancer radiotherapy.

    Directory of Open Access Journals (Sweden)

    Qinquan Gao

    Full Text Available The combination external-beam radiotherapy and high-dose-rate brachytherapy is a standard form of treatment for patients with locally advanced uterine cervical cancer. Personalized radiotherapy in cervical cancer requires efficient and accurate dose planning and assessment across these types of treatment. To achieve radiation dose assessment, accurate mapping of the dose distribution from HDR-BT onto EBRT is extremely important. However, few systems can achieve robust dose fusion and determine the accumulated dose distribution during the entire course of treatment. We have therefore developed a toolbox (FZUImageReg, which is a user-friendly dose fusion system based on hybrid image registration for radiation dose assessment in cervical cancer radiotherapy. The main part of the software consists of a collection of medical image registration algorithms and a modular design with a user-friendly interface, which allows users to quickly configure, test, monitor, and compare different registration methods for a specific application. Owing to the large deformation, the direct application of conventional state-of-the-art image registration methods is not sufficient for the accurate alignment of EBRT and HDR-BT images. To solve this problem, a multi-phase non-rigid registration method using local landmark-based free-form deformation is proposed for locally large deformation between EBRT and HDR-BT images, followed by intensity-based free-form deformation. With the transformation, the software also provides a dose mapping function according to the deformation field. The total dose distribution during the entire course of treatment can then be presented. Experimental results clearly show that the proposed system can achieve accurate registration between EBRT and HDR-BT images and provide radiation dose warping and fusion results for dose assessment in cervical cancer radiotherapy in terms of high accuracy and efficiency.

  1. Determinación de la distribución de dosis en tratamientos de cáncer de mama con fuentes de 192 Ir HDR / Determination of doses distributions on breast cancer treatments with 192 Ir HDR sources

    OpenAIRE

    Beltrán Gómez, Cristian Camilo

    2010-01-01

    La Braquiterapia de alta tasa de dosis HDR es una moderna técnica de tratamiento que ha venido teniendo un rápido crecimiento en su uso clínico, reemplazando a la braquiterapia de baja tasa de dosis LDR. La braquiterapia HDR se caracteriza por utilizar fuentes radiactivas con tasas de dosis mayores a 12 Gy/h, por tanto con propósitos de protección radiológica debe ser realizada con equipos de carga remota, es común en los tratamientos de braquiterapia HDR depositar altas dosis por fracción, c...

  2. Prostate-Specific Antigen Bounce After High-Dose-Rate Monotherapy for Prostate Cancer

    International Nuclear Information System (INIS)

    Mehta, Niraj H.; Kamrava, Mitchell; Wang, Pin-Chieh; Steinberg, Michael; Demanes, Jeffrey

    2013-01-01

    Purpose: To characterize the magnitude and kinetics of prostate-specific antigen (PSA) bounces after high-dose-rate (HDR) monotherapy and determine relationships between certain clinical factors and PSA bounce. Methods and Materials: Longitudinal PSA data and various clinical parameters were examined in 157 consecutive patients treated with HDR monotherapy between 1996 and 2005. We used the following definition for PSA bounce: rise in PSA ≥threshold, after which it returns to the prior level or lower. Prostate-specific antigen failure was defined per the Phoenix definition (nadir +2 ng/mL). Results: A PSA bounce was noted in 67 patients (43%). The number of bounces per patient was 1 in 45 cases (67%), 2 in 19 (28%), 3 in 2 (3%), 4 in 0, and 5 in 1 (1%). The median time to maximum PSA bounce was 1.3 years, its median magnitude was 0.7, and its median duration was 0.75 years. Three patients (2%) were noted to have PSA failure. None of the 3 patients who experienced biochemical failure exhibited PSA bounce. In the fully adjusted model for predicting each bounce, patients aged <55 years had a statistically significant higher likelihood of experiencing a bounce (odds ratio 2.22, 95% confidence interval 1.38-3.57, P=.001). There was also a statistically significant higher probability of experiencing a bounce for every unit decrease in Gleason score (odds ratio 1.52, 95% confidence interval 1.01-2.04, P=.045). Conclusions: A PSA bounce occurs in a significant percentage of patients treated with HDR monotherapy, with magnitudes varying from <1 in 28% of cases to ≥1 in 15%. The median duration of bounce is <1 year. More bounces were identified in patients with lower Gleason score and age <55 years. Further investigation using a model to correlate magnitude and frequency of bounces with clinical variables are under way

  3. HDR Efex Pro After the Shoot

    CERN Document Server

    Sholik, Stan

    2011-01-01

    A concise, on-the-go guide to the new HDR Efex Pro imaging toolkit for photographers Now that you've gone mobile and HDR, you want to be able to download and enhance your favorite photos on the run, without having to return to the mother ship (i.e., your desktop computer). This book shows you just how to do that using the amazing HDR Efex Pro, the image editing toolset from Nik Software. In brilliant color and using plenty of show-stopping examples, this practical guide explains all tools and features. Follow numbered steps and you'll soon be handling things like alignment, ghosting control, h

  4. Cost modelling of electricity-producing hot dry rock (HDR) geothermal systems in the United Kingdom

    International Nuclear Information System (INIS)

    Doherty, P.; Harrison, R.

    1995-01-01

    A detailed and comprehensive cost model for Hot Dry Rock (HDR) electricity producing systems has been developed in this study. The model takes account of the major aspects of the HDR system, parameterized in terms of the main physical and cost parameters of the resource and the utilization system. A doublet configuration is assumed, and the conceptual HDR system which is defined in the study is based upon the UK Department of Energy (DEn) HDR geothermal R and D programme. The model has been used to calculate the costs of HDR electricity for a UK defined base case which represents a consensus view of what might be achieved in Cornwall in the long term. At 14.2 p/kWh (1988 costs) this cost appears to be unacceptably high. A wide-ranging sensitivity study has also been carried out on the main resource, geometrical, and operational parameters of the HDR system centred around the UK base case. The sensitivity study shows the most important parameters to be thermal gradient and depth. The geometrical arrangement and the shape of the reservoir constitute major uncertainties in HDR systems. Their effect on temperature has a major influence on system performance, and therefore a range of theoretically possible geometries have been studied and the importance of geometrical effects on HDR electricity costs assessed. The most cost effective HDR arrangement in terms of optimized volumes and flow rates has been investigated for a world-wide range of thermal settings. The main conclusions from this study suggests that for HDR electricity to be economic, thermal gradients of 55 o C/km and above, well depths of 5 km or less, and production fluid temperatures of 210 o C and above are required. (UK)

  5. Adaptive error detection for HDR/PDR brachytherapy: Guidance for decision making during real-time in vivo point dosimetry

    DEFF Research Database (Denmark)

    Kertzscher Schwencke, Gustavo Adolfo Vladimir; Andersen, Claus E.; Tanderup, Kari

    2014-01-01

    Purpose:This study presents an adaptive error detection algorithm (AEDA) for real-timein vivo point dosimetry during high dose rate (HDR) or pulsed dose rate (PDR) brachytherapy (BT) where the error identification, in contrast to existing approaches, does not depend on an a priori reconstruction ......, and the AEDA’s capacity to distinguish between true and false error scenarios. The study further shows that the AEDA can offer guidance in decision making in the event of potential errors detected with real-time in vivo point dosimetry....... of the dosimeter position reconstruction. Given its nearly exclusive dependence on stable dosimeter positioning, the AEDA allows for a substantially simplified and time efficient real-time in vivo BT dosimetry implementation. Methods:In the event of a measured potential treatment error, the AEDA proposes the most...

  6. The influence of the dwell time deviation constraint (DTDC) parameter on dosimetry with IPSA optimisation for HDR prostate brachytherapy

    International Nuclear Information System (INIS)

    Smith, Ryan L.; Millar, Jeremy L.; Panettieri, Vanessa; Mason, Natasha; Lancaster, Craig; Francih, Rick D.

    2015-01-01

    To investigate how the dwell time deviation constraint (DTDC) parameter, applied to inverse planning by simulated annealing (IPSA) optimisation limits large dwell times from occurring in each catheter and to characterise the effect on the resulting dosimetry for prostate high dose rate (HDR) brachytherapy treatment plans. An unconstrained IPSA optimised treatment plan, using the Oncentra Brachytherapy treatment planning system (version 4.3, Nucletron an Elekta company, Elekta AB, Stockholm, Sweden), was generated for 20 consecutive HDR prostate brachytherapy patients, with the DTDC set to zero. Successive constrained optimisation plans were also created for each patient by increasing the DTDC parameter by 0.2, up to a maximum value of 1.0. We defined a “plan modulation indexâ€, to characterise the change of dwell time modulation as the DTDC parameter was increased. We calculated the dose volume histogram indices for the PTV (D90, V100, V150, V200%) and urethra (D10%) to characterise the effect on the resulting dosimetry. The average PTV D90% decreases as the DTDC is applied, on average by only 1.5 %, for a DTDC = 0.4. The measures of high dose regions in the PTV, V150 and V200%, increase on average by less than 5 and 2 % respectively. The net effect of DTDC on the modulation of dwell times has been characterised by the introduction of the plan modulation index. DTDC applied during IPSA optimisation of HDR prostate brachytherapy plans reduce the occurrence of large isolated dwell times within individual catheters. The mechanism by which DTDC works has been described and its effect on the modulation of dwell times has been characterised. The authors recommend using a DTDC parameter no greater than 0.4 to obtain a plan with dwell time modulation comparable to a geometric optimised plan. This yielded on average a 1.5 % decrease in PTV coverage and an acceptable increase in V150%, without compromising the urethral dose.

  7. HDR economy with special reference to conditions in Europe

    International Nuclear Information System (INIS)

    Smolka, K.; Kappelmeyer, O.

    1992-01-01

    A cost benefit model for the economic evaluation of HDR energy production cost was developed. It can be used for systematic cost analyses of different components of a HDR plant. This relates to natural subsurface conditions (i.e., geothermal gradient, tectonic stress) as well as technical components (i.e., boreholes, heat/power conversion system). The model provides the limiting conditions for an economic HDR energy production. The economic model was applied for an evaluation of the natural parameters in the subsurface of Germany regarding their HDR-suitability. Within a site selection program in Europe the model is part of feasibility studies for the conceptual design of a HDR demonstration plant at three candidate sites: Bad Urach (Germany); Soultz-sous-Forets (France) and Cornwall (UK)

  8. An Eight-Year Experience of HDR Brachytherapy Boost for Localized Prostate Cancer: Biopsy and PSA Outcome

    International Nuclear Information System (INIS)

    Bachand, Francois; Martin, Andre-Guy; Beaulieu, Luc; Harel, Francois M.Sc.; Vigneault, Eric

    2009-01-01

    Purpose: To evaluate the biochemical recurrence-free survival (bRFS), the 2-year biopsy outcome and the prostate-specific antigen (PSA) bounce in patients with localized prostate cancer treated with an inversely planned high-dose-rate (HDR) brachytherapy boost. Materials and methods: Data were collected from 153 patients treated between 1999 and 2006 with external beam pelvic radiation followed by an HDR Ir-192 prostate boost. These patients were given a boost of 18 to 20 Gy using inverse-planning with simulated annealing (IPSA).We reviewed and analyzed all prostate-specific antigen levels and control biopsies. Results: The median follow-up was 44 months (18-95 months). When categorized by risk of progression, 74.5% of patients presented an intermediate risk and 14.4% a high one. Prostate biopsies at 2 years posttreatment were negative in 86 of 94 patients (91.5%), whereas two biopsies were inconclusive. Biochemical control at 60 months was at 96% according to the American Society for Therapeutic Radiology and Oncology and the Phoenix consensus definitions. A PSA bounce (PSA values of 2 ng/mL or more above nadir) was observed in 15 patients of 123 (9.8%). The median time to bounce was 15.2 months (interquartile range, 11.0-17.7) and the median bounce duration 18.7 months (interquartile range, 12.1-29). The estimate of overall survival at 60 months was 97.1% (95% CI, 91.6-103%). Conclusions: Considering that inverse planned HDR brachytherapy prostate boosts led to an excellent biochemical response, with a 2-year negative biopsy rate, we recommend a conservative approach in face of a PSA bounce even though it was observed in 10% of patients

  9. TU-H-CAMPUS-JeP3-05: Adaptive Determination of Needle Sequence HDR Prostate Brachytherapy with Divergent Needle-By-Needle Delivery

    International Nuclear Information System (INIS)

    Borot de Battisti, M; Maenhout, M; Lagendijk, J J W; Van Vulpen, M; Moerland, M A; Denis de Senneville, B; Hautvast, G; Binnekamp, D

    2016-01-01

    Purpose: To develop a new method which adaptively determines the optimal needle insertion sequence for HDR prostate brachytherapy involving divergent needle-by-needle dose delivery by e.g. a robotic device. A needle insertion sequence is calculated at the beginning of the intervention and updated after each needle insertion with feedback on needle positioning errors. Methods: Needle positioning errors and anatomy changes may occur during HDR brachytherapy which can lead to errors in the delivered dose. A novel strategy was developed to calculate and update the needle sequence and the dose plan after each needle insertion with feedback on needle positioning errors. The dose plan optimization was performed by numerical simulations. The proposed needle sequence determination optimizes the final dose distribution based on the dose coverage impact of each needle. This impact is predicted stochastically by needle insertion simulations. HDR procedures were simulated with varying number of needle insertions (4 to 12) using 11 patient MR data-sets with PTV, prostate, urethra, bladder and rectum delineated. Needle positioning errors were modeled by random normally distributed angulation errors (standard deviation of 3 mm at the needle’s tip). The final dose parameters were compared in the situations where the needle with the largest vs. the smallest dose coverage impact was selected at each insertion. Results: Over all scenarios, the percentage of clinically acceptable final dose distribution improved when the needle selected had the largest dose coverage impact (91%) compared to the smallest (88%). The differences were larger for few (4 to 6) needle insertions (maximum difference scenario: 79% vs. 60%). The computation time of the needle sequence optimization was below 60s. Conclusion: A new adaptive needle sequence determination for HDR prostate brachytherapy was developed. Coupled to adaptive planning, the selection of the needle with the largest dose coverage impact

  10. Evaluation of water-mimicking solid phantom materials for use in HDR and LDR brachytherapy dosimetry

    Science.gov (United States)

    Schoenfeld, Andreas A.; Thieben, Maike; Harder, Dietrich; Poppe, Björn; Chofor, Ndimofor

    2017-12-01

    In modern HDR or LDR brachytherapy with photon emitters, fast checks of the dose profiles generated in water or a water-equivalent phantom have to be available in the interest of patient safety. However, the commercially available brachytherapy photon sources cover a wide range of photon emission spectra, and the range of the in-phantom photon spectrum is further widened by Compton scattering, so that the achievement of water-mimicking properties of such phantoms involves high requirements on their atomic composition. In order to classify the degree of water equivalence of the numerous commercially available solid water-mimicking phantom materials and the energy ranges of their applicability, the radial profiles of the absorbed dose to water, D w, have been calculated using Monte Carlo simulations in these materials and in water phantoms of the same dimensions. This study includes the HDR therapy sources Nucletron Flexisource Co-60 HDR (60Co), Eckert und Ziegler BEBIG GmbH CSM-11 (137Cs), Implant Sciences Corporation HDR Yb-169 Source 4140 (169Yb) as well as the LDR therapy sources IsoRay Inc. Proxcelan CS-1 (131Cs), IsoAid Advantage I-125 IAI-125A (125I), and IsoAid Advantage Pd-103 IAPd-103A (103Pd). Thereby our previous comparison between phantom materials and water surrounding a Varian GammaMed Plus HDR therapy 192Ir source (Schoenfeld et al 2015) has been complemented. Simulations were performed in cylindrical phantoms consisting of either water or the materials RW1, RW3, Solid Water, HE Solid Water, Virtual Water, Plastic Water DT, Plastic Water LR, Original Plastic Water (2015), Plastic Water (1995), Blue Water, polyethylene, polystyrene and PMMA. While for 192Ir, 137Cs and 60Co most phantom materials can be regarded as water equivalent, for 169Yb the materials Plastic Water LR, Plastic Water DT and RW1 appear as water equivalent. For the low-energy sources 106Pd, 131Cs and 125I, only Plastic Water LR can be classified as water equivalent.

  11. Feasibility and early outcome of high-dose-rate Ir-192 brachytherapy as monotherapy in two fractions within 1 day for high-/very high-risk prostate cancer.

    Science.gov (United States)

    Ashida, Shingo; Yamasaki, Ichiro; Tamura, Kenji; Shimamoto, Tsutomu; Inoue, Keiji; Kariya, Shinji; Kobayashi, Kana; Yamagami, Takuji; Shuin, Taro

    2016-05-01

    The aim of the present study was to evaluate the feasibility and preliminary outcomes of high-dose-rate (HDR)-brachytherapy as a monotherapy in two fractions within 1 day for localized prostate cancer, including high-/very high-risk cases. Among the 68 patients treated with HDR monotherapy between July 2011 and December 2014, 65 had a minimal follow-up of 12 months without adjuvant androgen deprivation therapy and were enrolled in the present study [42/65 (64.6%) exhibited high-/very high-risk diseases]. HDR monotherapy was performed in two fractions with a minimal interval of 6 h and the prescribed dose was 13.5 Gy (×2). Adverse events (AEs) were assessed using Common Terminology Criteria for Adverse Events (version 4; http://ctep.cancer.gov/protocolDevelopment/electronic_applications/ctc.htm#ctc_40), and biochemical failure was assessed by the Phoenix definition. The median follow-up time was 30.1 months. The majority of patients had Grade 0-1 acute AEs. Four patients (6.2%) exhibited urinary retention, requiring a Foley catheter. Grade 3 acute AEs occurred at a frequency of 3.1% and hematuria at 1.5%. The majority of patients also exhibited Grade 0-1 chronic AEs. Grade 3 chronic AEs occurred at a frequency of 1.5% and urethral stricture at 1.5%, for which endoscopic treatment was indicated. Acute and chronic gastrointestinal AEs were uncommon, and no Grade 3 or above AEs developed. Biochemical failure occurred in 4 patients who all exhibited high-/very high-risk diseases. Kaplan-Meier estimated that 3 year biochemical failure-free survival was 91.6% overall and 88.0% in high-/very high-risk cases. The present two-fraction 1 day HDR monotherapy is feasible with minimal AEs and achieved acceptable biochemical control of localized prostate cancer, including high-/very high-risk cases, although long-term follow-up is required.

  12. Comparison between beta radiation dose distribution due to LDR and HDR ocular brachytherapy applicators using GATE Monte Carlo platform.

    Science.gov (United States)

    Mostafa, Laoues; Rachid, Khelifi; Ahmed, Sidi Moussa

    2016-08-01

    Eye applicators with 90Sr/90Y and 106Ru/106Rh beta-ray sources are generally used in brachytherapy for the treatment of eye diseases as uveal melanoma. Whenever, radiation is used in treatment, dosimetry is essential. However, knowledge of the exact dose distribution is a critical decision-making to the outcome of the treatment. The Monte Carlo technique provides a powerful tool for calculation of the dose and dose distributions which helps to predict and determine the doses from different shapes of various types of eye applicators more accurately. The aim of this work consisted in using the Monte Carlo GATE platform to calculate the 3D dose distribution on a mathematical model of the human eye according to international recommendations. Mathematical models were developed for four ophthalmic applicators, two HDR 90Sr applicators SIA.20 and SIA.6, and two LDR 106Ru applicators, a concave CCB model and a flat CCB model. In present work, considering a heterogeneous eye phantom and the chosen tumor, obtained results with the use of GATE for mean doses distributions in a phantom and according to international recommendations show a discrepancy with respect to those specified by the manufacturers. The QC of dosimetric parameters shows that contrarily to the other applicators, the SIA.20 applicator is consistent with recommendations. The GATE platform show that the SIA.20 applicator present better results, namely the dose delivered to critical structures were lower compared to those obtained for the other applicators, and the SIA.6 applicator, simulated with MCNPX generates higher lens doses than those generated by GATE. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  13. Interactive multiobjective optimization for anatomy-based three-dimensional HDR brachytherapy

    Science.gov (United States)

    Ruotsalainen, Henri; Miettinen, Kaisa; Palmgren, Jan-Erik; Lahtinen, Tapani

    2010-08-01

    In this paper, we present an anatomy-based three-dimensional dose optimization approach for HDR brachytherapy using interactive multiobjective optimization (IMOO). In brachytherapy, the goals are to irradiate a tumor without causing damage to healthy tissue. These goals are often conflicting, i.e. when one target is optimized the other will suffer, and the solution is a compromise between them. IMOO is capable of handling multiple and strongly conflicting objectives in a convenient way. With the IMOO approach, a treatment planner's knowledge is used to direct the optimization process. Thus, the weaknesses of widely used optimization techniques (e.g. defining weights, computational burden and trial-and-error planning) can be avoided, planning times can be shortened and the number of solutions to be calculated is small. Further, plan quality can be improved by finding advantageous trade-offs between the solutions. In addition, our approach offers an easy way to navigate among the obtained Pareto optimal solutions (i.e. different treatment plans). When considering a simulation model of clinical 3D HDR brachytherapy, the number of variables is significantly smaller compared to IMRT, for example. Thus, when solving the model, the CPU time is relatively short. This makes it possible to exploit IMOO to solve a 3D HDR brachytherapy optimization problem. To demonstrate the advantages of IMOO, two clinical examples of optimizing a gynecologic cervix cancer treatment plan are presented.

  14. Interactive multiobjective optimization for anatomy-based three-dimensional HDR brachytherapy

    International Nuclear Information System (INIS)

    Ruotsalainen, Henri; Miettinen, Kaisa; Palmgren, Jan-Erik; Lahtinen, Tapani

    2010-01-01

    In this paper, we present an anatomy-based three-dimensional dose optimization approach for HDR brachytherapy using interactive multiobjective optimization (IMOO). In brachytherapy, the goals are to irradiate a tumor without causing damage to healthy tissue. These goals are often conflicting, i.e. when one target is optimized the other will suffer, and the solution is a compromise between them. IMOO is capable of handling multiple and strongly conflicting objectives in a convenient way. With the IMOO approach, a treatment planner's knowledge is used to direct the optimization process. Thus, the weaknesses of widely used optimization techniques (e.g. defining weights, computational burden and trial-and-error planning) can be avoided, planning times can be shortened and the number of solutions to be calculated is small. Further, plan quality can be improved by finding advantageous trade-offs between the solutions. In addition, our approach offers an easy way to navigate among the obtained Pareto optimal solutions (i.e. different treatment plans). When considering a simulation model of clinical 3D HDR brachytherapy, the number of variables is significantly smaller compared to IMRT, for example. Thus, when solving the model, the CPU time is relatively short. This makes it possible to exploit IMOO to solve a 3D HDR brachytherapy optimization problem. To demonstrate the advantages of IMOO, two clinical examples of optimizing a gynecologic cervix cancer treatment plan are presented.

  15. High-dose-rate brachytherapy as monotherapy delivered in two fractions within one day for favorable/intermediate-risk prostate cancer: preliminary toxicity data.

    Science.gov (United States)

    Ghilezan, Michel; Martinez, Alvaro; Gustason, Gary; Krauss, Daniel; Antonucci, J Vito; Chen, Peter; Fontanesi, James; Wallace, Michelle; Ye, Hong; Casey, Alyse; Sebastian, Evelyn; Kim, Leonard; Limbacher, Amy

    2012-07-01

    To report the toxicity profile of high-dose-rate (HDR)-brachytherapy (BT) as monotherapy in a Human Investigation Committee-approved study consisting of a single implant and two fractions (12 Gy × 2) for a total dose of 24 Gy, delivered within 1 day. The dose was subsequently increased to 27 Gy (13.5 Gy × 2) delivered in 1 day. We report the acute and early chronic genitourinary and gastrointestinal toxicity. A total of 173 patients were treated between December 2005 and July 2010. However, only the first 100 were part of the IRB-approved study and out of these, only 94 had a minimal follow-up of 6 months, representing the study population for this preliminary report. All patients had clinical Stage T2b or less (American Joint Committee on Cancer, 5th edition), Gleason score 6-7 (3+4), and prostate-specific antigen level of ≤12 ng/mL. Ultrasound-guided HDR-BT with real-time dosimetry was used. The prescription dose was 24 Gy for the first 50 patients and 27 Gy thereafter. The dosimetric goals and constraints were the same for the two dose groups. Toxicity was scored using the National Cancer Institute Common Terminology Criteria for Adverse Events, version 3. The highest toxicity scores encountered at any point during follow-up are reported. The median follow-up was 17 months (range, 6-40.5). Most patients had Grade 0-1 acute toxicity. The Grade 2 acute genitourinary toxicity was mainly frequency/urgency (13%), dysuria (5%), hematuria, and dribbling/hesitancy (2%). None of the patients required a Foley catheter at any time; however, 8% of the patients experienced transient Grade 1 diarrhea. No other acute gastrointestinal toxicities were found. The most common chronic toxicity was Grade 2 urinary frequency/urgency in 16% of patients followed by dysuria in 4% of patients; 2 patients had Grade 2 rectal bleeding and 1 had Grade 4, requiring laser treatment. Favorable-risk prostate cancer patients treated with a single implant HDR-BT to 24-27 Gy in two fractions

  16. High-Dose-Rate Brachytherapy as Monotherapy Delivered in Two Fractions Within One Day for Favorable/Intermediate-Risk Prostate Cancer: Preliminary Toxicity Data

    Energy Technology Data Exchange (ETDEWEB)

    Ghilezan, Michel, E-mail: mghilezan@beaumont.edu [Department of Radiation Oncology, William Beaumont Hospital and Rose Cancer Institute, Royal Oak, Michigan (United States); Martinez, Alvaro; Gustason, Gary; Krauss, Daniel; Antonucci, J. Vito; Chen, Peter; Fontanesi, James; Wallace, Michelle; Ye Hong; Casey, Alyse; Sebastian, Evelyn; Kim, Leonard; Limbacher, Amy [Department of Radiation Oncology, William Beaumont Hospital and Rose Cancer Institute, Royal Oak, Michigan (United States)

    2012-07-01

    Purpose: To report the toxicity profile of high-dose-rate (HDR)-brachytherapy (BT) as monotherapy in a Human Investigation Committee-approved study consisting of a single implant and two fractions (12 Gy Multiplication-Sign 2) for a total dose of 24 Gy, delivered within 1 day. The dose was subsequently increased to 27 Gy (13.5 Gy Multiplication-Sign 2) delivered in 1 day. We report the acute and early chronic genitourinary and gastrointestinal toxicity. Methods and Materials: A total of 173 patients were treated between December 2005 and July 2010. However, only the first 100 were part of the IRB-approved study and out of these, only 94 had a minimal follow-up of 6 months, representing the study population for this preliminary report. All patients had clinical Stage T2b or less (American Joint Committee on Cancer, 5th edition), Gleason score 6-7 (3+4), and prostate-specific antigen level of {<=}12 ng/mL. Ultrasound-guided HDR-BT with real-time dosimetry was used. The prescription dose was 24 Gy for the first 50 patients and 27 Gy thereafter. The dosimetric goals and constraints were the same for the two dose groups. Toxicity was scored using the National Cancer Institute Common Terminology Criteria for Adverse Events, version 3. The highest toxicity scores encountered at any point during follow-up are reported. Results: The median follow-up was 17 months (range, 6-40.5). Most patients had Grade 0-1 acute toxicity. The Grade 2 acute genitourinary toxicity was mainly frequency/urgency (13%), dysuria (5%), hematuria, and dribbling/hesitancy (2%). None of the patients required a Foley catheter at any time; however, 8% of the patients experienced transient Grade 1 diarrhea. No other acute gastrointestinal toxicities were found. The most common chronic toxicity was Grade 2 urinary frequency/urgency in 16% of patients followed by dysuria in 4% of patients; 2 patients had Grade 2 rectal bleeding and 1 had Grade 4, requiring laser treatment. Conclusions: Favorable

  17. High-Dose-Rate Brachytherapy as Monotherapy Delivered in Two Fractions Within One Day for Favorable/Intermediate-Risk Prostate Cancer: Preliminary Toxicity Data

    International Nuclear Information System (INIS)

    Ghilezan, Michel; Martinez, Alvaro; Gustason, Gary; Krauss, Daniel; Antonucci, J. Vito; Chen, Peter; Fontanesi, James; Wallace, Michelle; Ye Hong; Casey, Alyse; Sebastian, Evelyn; Kim, Leonard; Limbacher, Amy

    2012-01-01

    Purpose: To report the toxicity profile of high-dose-rate (HDR)-brachytherapy (BT) as monotherapy in a Human Investigation Committee-approved study consisting of a single implant and two fractions (12 Gy × 2) for a total dose of 24 Gy, delivered within 1 day. The dose was subsequently increased to 27 Gy (13.5 Gy × 2) delivered in 1 day. We report the acute and early chronic genitourinary and gastrointestinal toxicity. Methods and Materials: A total of 173 patients were treated between December 2005 and July 2010. However, only the first 100 were part of the IRB-approved study and out of these, only 94 had a minimal follow-up of 6 months, representing the study population for this preliminary report. All patients had clinical Stage T2b or less (American Joint Committee on Cancer, 5th edition), Gleason score 6-7 (3+4), and prostate-specific antigen level of ≤12 ng/mL. Ultrasound-guided HDR-BT with real-time dosimetry was used. The prescription dose was 24 Gy for the first 50 patients and 27 Gy thereafter. The dosimetric goals and constraints were the same for the two dose groups. Toxicity was scored using the National Cancer Institute Common Terminology Criteria for Adverse Events, version 3. The highest toxicity scores encountered at any point during follow-up are reported. Results: The median follow-up was 17 months (range, 6–40.5). Most patients had Grade 0-1 acute toxicity. The Grade 2 acute genitourinary toxicity was mainly frequency/urgency (13%), dysuria (5%), hematuria, and dribbling/hesitancy (2%). None of the patients required a Foley catheter at any time; however, 8% of the patients experienced transient Grade 1 diarrhea. No other acute gastrointestinal toxicities were found. The most common chronic toxicity was Grade 2 urinary frequency/urgency in 16% of patients followed by dysuria in 4% of patients; 2 patients had Grade 2 rectal bleeding and 1 had Grade 4, requiring laser treatment. Conclusions: Favorable-risk prostate cancer patients treated with

  18. TU-C-201-02: Clinical Implementation of HDR: Afterloader and Applicator Selection

    Energy Technology Data Exchange (ETDEWEB)

    Esthappan, J. [Washington University School of Medicine (United States)

    2015-06-15

    Recent use of HDR has increased while planning has become more complex often necessitating 3D image-based planning. While many guidelines for the use of HDR exist, they have not kept pace with the increased complexity of 3D image-based planning. Furthermore, no comprehensive document exists to describe the wide variety of current HDR clinical indications. This educational session aims to summarize existing national and international guidelines for the safe implementation of an HDR program. A summary of HDR afterloaders available on the market and their existing applicators will be provided, with guidance on how to select the best fit for each institution’s needs. Finally, the use of checklists will be discussed as a means to implement a safe and efficient HDR program and as a method by which to verify the quality of an existing HDR program. This session will provide the perspective of expert HDR physicists as well as the perspective of a new HDR user. Learning Objectives: Summarize national and international safety and staffing guidelines for HDR implementation Discuss the process of afterloader and applicator selection for gynecologic, prostate, breast, interstitial, surface treatments Learn about the use of an audit checklist tool to measure of quality control of a new or existing HDR program Describe the evolving use of checklists within an HDR program.

  19. Comparison of absorbed dose in the cervix carcinoma therapy by brachytherapy of high dose rate using the conventional planning and Monte Carlo simulation; Comparacao da dose absorvida no tratamento do cancer ginecologico por braquiterapia de alta taxa de dose utilizando o planejamento convencional do tratamento e simulacao de Monte Carlo

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Aneli Oliveira da

    2010-07-01

    This study aims to compare the doses received for patients submitted to brachytherapy High Dose Rate (HDR) brachytherapy, a method of treatment of the cervix carcinoma, performed in the planning system PLATO BPS with the doses obtained by Monte Carlo simulation using the radiation transport code MCNP 5 and one female anthropomorphic phantom based on voxel, the FAX. The implementation of HDR brachytherapy treatment for the cervix carcinoma consists of the insertion of an intrauterine probe and an intravaginal probe (ring or ovoid) and then two radiographs are obtained, anteroposterior (AP) and lateral (LAT) to confirm the position of the applicators in the patient and to allow the treatment planning and the determination of the absorbed dose at points of interest: rectum, bladder, sigmoid and point A, which corresponds anatomically to the crossings of the uterine arteries with ureters The absorbed doses obtained with the code MCNP 5, with the exception of the absorbed dose in the rectum and sigmoid for the simulation considering a point source of {sup 192}Ir, are lower than the absorbed doses from PLATO BPS calculations because the MCNP 5 considers the chemical compositions and densities of FAX body, not considering the medium as water. When considering the Monte Carlo simulation for a source with dimensions equal to that used in the brachytherapy irradiator used in this study, the values of calculated absorbed dose to the bladder, to the rectum, to the right point A and to the left point A were respectively lower than those determined by the treatment planning system in 33.29, 5.01, 22.93 and 19.04%. These values are almost all larger than the maximum acceptable deviation between patient planned and administered doses (5 %). With regard to the rectum and bladder, which are organs that must be protected, the present results are in favor of the radiological protection of patients. The point A, that is on the isodose of 100%, used to tumor treatment, the results

  20. Commissioning and quality assurance procedures for the HDR Valencia skin applicators

    Directory of Open Access Journals (Sweden)

    Domingo Granero

    2016-11-01

    Full Text Available The Valencia applicators (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden are cup-shaped tungsten applicators with a flattening filter used to collimate the radiation produced by a high-dose-rate (HDR 192 Ir source, and provide a homogeneous absorbed dose at a given depth. This beam quality provides a good option for the treatment of skin lesions at shallow depth (3-4 mm. The user must perform commissioning and periodic testing of these applicators to guarantee the proper and safe delivery of the intended absorbed dose, as recommended in the standards in radiation oncology. In this study, based on AAPM and GEC-ESTRO guidelines for brachytherapy units and our experience, a set of tests for the commissioning and periodic testing of the Valencia applicators is proposed. These include general considerations, verification of the manufacturer documentation and physical integrity, evaluation of the source-to-indexer distance and reproducibility, setting the library plan in the treatment planning system, evaluation of flatness and symmetry, absolute output and percentage depth dose verification, independent calculation of the treatment time, and visual inspection of the applicator before each treatment. For each test, the proposed methodology, equipment, frequency, expected results, and tolerance levels (when applicable are provided.

  1. Mixed integer programming improves comprehensibility and plan quality in inverse optimization of prostate HDR Brachytherapy

    NARCIS (Netherlands)

    Gorissen, B.L.; den Hertog, D.; Hoffmann, A.L.

    2013-01-01

    Current inverse treatment planning methods that optimize both catheter positions and dwell times in prostate HDR brachytherapy use surrogate linear or quadratic objective functions that have no direct interpretation in terms of dose-volume histogram (DVH) criteria, do not result in an optimum or

  2. Prospective Multi-Institutional Study of Definitive Radiotherapy With High-Dose-Rate Intracavitary Brachytherapy in Patients With Nonbulky (<4-cm) Stage I and II Uterine Cervical Cancer (JAROG0401/JROSG04-2)

    Energy Technology Data Exchange (ETDEWEB)

    Toita, Takafumi, E-mail: b983255@med.u-ryukyu.ac.jp [Department of Radiology, Graduate School of Medical Science, University of Ryukyus, Okinawa (Japan); Kato, Shingo [Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba (Japan); Niibe, Yuzuru [Department of Radiology, School of Medicine, Kitasato University, Sagamihara (Japan); Ohno, Tatsuya [Gunma University Heavy Ion Medical Center, Maebashi (Japan); Kazumoto, Tomoko [Department of Radiology, Saitama Cancer Center, Saitama (Japan); Kodaira, Takeshi [Department of Radiation Oncology, Aichi Cancer Center, Nagoya (Japan); Kataoka, Masaaki [Department of Radiology, National Shikoku Cancer Center, Ehime (Japan); Shikama, Naoto [Department of Radiation Oncology, Saku Central Hospital, Saku (Japan); Kenjo, Masahiro [Department of Radiation Oncology, Graduate School of Medical Science, Hiroshima University, Hiroshima (Japan); Tokumaru, Sunao [Department of Radiology, Saga University, Saga (Japan); Yamauchi, Chikako [Department of Radiation Oncology, Shiga Medical Center for Adults, Moriyama (Japan); Suzuki, Osamu [Department of Radiation Oncology, Osaka Medical Center for Cancer, Osaka (Japan); Sakurai, Hideyuki [Proton Medical Research Center and Tsukuba University, Tsukuba (Japan); Numasaki, Hodaka; Teshima, Teruki [Department of Medical Physics and Engineering, Graduate School of Medicine, Osaka University, Suita, Osaka (Japan); Oguchi, Masahiko [Department of Radiation Oncology, Cancer Institute Hospital, Tokyo (Japan); Kagami, Yoshikazu [Radiation Oncology Division, National Cancer Center Hospital, Tokyo (Japan); Nakano, Takashi [Department of Radiation Oncology, Gunma University, Graduate School of Medicine, Maebashi (Japan); Hiraoka, Masahiro [Department of Radiation Oncology and Image-applied Therapy, Kyoto University, Graduate School of Medicine, Kyoto (Japan); Mitsuhashi, Norio [Department of Radiation Oncology, Tokyo Women' s Medical University, Tokyo (Japan)

    2012-01-01

    Purpose: To determine the efficacy of a definitive radiotherapy protocol using high-dose-rate intracavitary brachytherapy (HDR-ICBT) with a low cumulative dose schedule in nonbulky early-stage cervical cancer patients, we conducted a prospective multi-institutional study. Methods and Materials: Eligible patients had squamous cell carcinoma of the intact uterine cervix, Federation of Gynecologic Oncology and Obstetrics (FIGO) stages Ib1, IIa, and IIb, tumor size <40 mm in diameter (assessed by T2-weighted magnetic resonance imaging), and no pelvic/para-aortic lymphadenopathy. The treatment protocol consisted of whole-pelvis external beam radiotherapy (EBRT) of 20 Gy/10 fractions, pelvic EBRT with midline block of 30 Gy/15 fractions, and HDR-ICBT of 24 Gy/4 fractions (at point A). The cumulative biologically effective dose (BED) was 62 Gy{sub 10} ({alpha}/{beta} = 10) at point A. The primary endpoint was the 2-year pelvic disease progression-free (PDPF) rate. All patients received a radiotherapy quality assurance review. Results: Between September 2004 and July 2007, 60 eligible patients were enrolled. Thirty-six patients were assessed with FIGO stage Ib1; 12 patients with stage IIa; and 12 patients with stage IIb. Median tumor diameter was 28 mm (range, 6-39 mm). Median overall treatment time was 43 days. Median follow-up was 49 months (range, 7-72 months). Seven patients developed recurrences: 3 patients had pelvic recurrences (2 central, 1 nodal), and 4 patients had distant metastases. The 2-year PDPF was 96% (95% confidence interval [CI], 92%-100%). The 2-year disease-free and overall survival rates were 90% (95% CI, 82%-98%) and 95% (95% CI, 89%-100%), respectively. The 2-year late complication rates (according to Radiation Therapy Oncology Group/European Organization for Research and Treatment of Cancer of Grade {>=}1) were 18% (95% CI, 8%-28%) for large intestine/rectum, 4% (95% CI, 0%-8%) for small intestine, and 0% for bladder. No Grade {>=}3 cases were

  3. Impact of oedema on implant geometry and dosimetry for temporary high dose rate brachytherapy of the prostate

    International Nuclear Information System (INIS)

    Kiffer, J.D.; Schumer, W.A.; Mantle, C.A.; McKenzie, B.J.; Feigen, M.; Quong, G.G.; Waterman, F.M.

    2003-01-01

    The optimal timing of dosimetry for permanent seed prostatic implants remains contentious given the half life of post-implant oedema resolution. The aim of this study was to establish whether prostatic oedematous change over the duration of a temporary high dose rate (HDR) interstitial brachytherapy (BR) boost would result in significant needle displacement, and whether this change in geometry would influence dosimetry. Two CT scans, one for dosimetric purposes on the day of the implant and the second just prior to implant removal, were obtained for four patients receiving transperineal interstitial prostate brachytherapy. The relative changes in cross-sectional dimensions of the implants were calculated by establishing the change in mean radial distance (MRD) of the needle positions from the geometric centre of the implant for each patient's pair of CT studies. The treatment plan, as calculated from the first CT scan, was used in the second set of CT images to allow a comparison of dose distribution. The percentage change in MRD over the duration of the temporary implants ranged from -1.91% to 1.95%. The maximum change in estimated volume was 3.94%. Dosimetric changes were negligible. In the four cases studied, the degree of oedematous change and consequent displacement of flexiguide needle positions was negligible and did not impact on the dosimetry. The rate and direction of oedematous change can be extremely variable but on the basis of the four cases studied and the results of a larger recent study, it might not be necessary to re-image patients for dosimetric purposes over the duration of a fractionated HDR BT boost to the prostate where flexiguide needles are utilized. Nevertheless, further investigation with larger patient numbers is required. Copyright (2003) Blackwell Science Pty Ltd

  4. SU-E-T-459: Impact of Source Position and Traveling Time On HDR Skin Surface Applicator Dosimetry

    International Nuclear Information System (INIS)

    Jeong, J; Barker, C; Zaider, M; Cohen, G

    2015-01-01

    Purpose: Observed dosimetric discrepancy between measured and treatment planning system (TPS) predicted values, during applicator commissioning, were traced to source position uncertainty in the applicator. We quantify the dosimetric impact of this geometric uncertainty, and of the source traveling time inside the applicator, and propose corrections for clinical use. Methods: We measured the dose profiles from the Varian Leipzig-style (horizontal) HDR skin applicator, using EBT3 film, photon diode, and optically stimulated luminescence dosimeter (OSLD) and three different GammaMed HDR afterloders. The dose profiles and depth dose of each aperture were measured at several depths (up to about 10 mm, depending on the dosimeter). The measured dose profiles were compared with Acuros calculated profiles in BrachyVision TPS. For the impact of the source position, EBT3 film measurements were performed with applicator, facing-down and facing-up orientations. The dose with and without source traveling was measured with diode detector using HDR timer and electrometer timer, respectively. Results: Depth doses measured using the three dosimeters were in good agreement, but were consistently higher than the Acuros dose calculations. Measurements with the applicator facing-up were significantly lower than those in the facing-down position with maximum difference of about 18% at the surface, due to source sag inside the applicator. Based on the inverse-square law, the effective source sag was evaluated to be about 0.5 mm from the planned position. The additional dose from the source traveling was about 2.8% for 30 seconds with 10 Ci source, decreasing with increased dwelling time and decreased source activity. Conclusion: Due to the short source-to-surface distance of the applicator, the small source sag inside the applicator has significant dosimetric impact, which should be considered before the clinical use of the applicator. Investigation of the effect for other applicators

  5. Ultrasound-guided high dose rate conformal brachytherapy boost in prostate cancer: treatment description and preliminary results of a phase I/II clinical trial

    Energy Technology Data Exchange (ETDEWEB)

    Stromberg, Jannifer; Martinez, Alvaro; Gonzalez, Jose; Edmundson, Gregory; Ohanian, Neshan; Vicini, Frank; Hollander, Jay; Gustafson, Gary; Spencer, William; Di, Yan; Brabbins, Donald

    1995-08-30

    Purpose: To improve results for locally advanced prostate cancer, a prospective clinical trial of concurrent external beam irradiation and fractionated iridium-192 (Ir-192) high dose rate (HDR) conformal boost brachytherapy was initiated. Methods and Materials: Between November 1991 and February 1994, 99 implants were performed on 33 patients with prostatic adenocarcinoma at William Beaumont Hospital. Using AJCC staging criteria, 9 patients had T2b tumors, 17 patients had T2c tumors, and 7 patients had T3 disease. Patients were treated with (a) 45.6 Gy whole pelvis external irradiation and (b) three HDR fractions of 5.5 Gy each (18 patients) or 6 Gy each (15 patients) to the prostate. Transperineal needle implants using real-time ultrasound guidance with interactive on-line isodose distributions were performed on an outpatient basis during weeks 1, 2, and 3 of external irradiation. Acute toxicity was scored using the Radiation Therapy Oncology Group (RTOG) morbidity grading system. Results: This technique of concurrent external pelvic irradiation and conformal HDR brachytherapy was well tolerated. No significant intraoperative or perioperative complications occurred. Three patients (9%) experienced Grade 3 acute toxicity (two dysuria and one diarrhea). All toxicities were otherwise Grades 1 or 2 and were primarily as expected from pelvic external irradiation. Persistent implant-related toxicities included Grades 1-2 perineal pain (12%) and hematospermia (15%). Median follow-up time was 13 months. Serum prostatic-specific antigen (PSA) levels normalized in 91% of patients (29 out of 32) within 1-14 months (median 2.8 months) after irradiation. PSA levels were progressively decreasing in the other three patients at last measurement. Prospectively planned prostatic rebiopsies done at 18 months in the first 10 patients were negative in 9 out of 10 (90%). Conclusions: Acute toxicity has been acceptable with this unique approach using conformal high dose rate Ir-192

  6. Endobronchial and endoesophageal high dose rate brachytherapy for malignant airway and digestive tract obstructions

    International Nuclear Information System (INIS)

    Mehta, Minesh P.

    1996-01-01

    With an annual incidence of more than 160,000 cases and a local failure rate between 30-50%, endobronchial occlusion seen with lung cancer is a common and potentially life-threatening complication. Several methods of managing this exist and recently endobronchial brachytherapy has been used extensively as a consequence of the development of fiberoptic bronchoscopy and high dose rate remote afterloading technology. Procedurally, one or more afterloading catheters are inserted in the involved portions of the tracheobronchial tree through fiberoptic guidance. Treatment techniques range from 1-4 applications fractionated over several weeks or given over 2 days with a single insertion procedure. Almost all procedures are currently performed in the outpatient setting. The major application of this technology is in the palliation of occlusive symptomatology. Clinical improvement ranges from 50-100%, radiographic reaeration ranges from 46-88% and bronchoscopic responses ranges from 59-100%. Symptomatic relief is usually quite durable with more than 70% of the patients' remaining life-time rendered symptom-free and symptom-improved. Recently, this modality has been explored for its curative potential as a boost following external beam radiotherapy. It is clear from these series, that in selected patients, endobronchial boost produces significant reaeration and sparing of lung volume from subsequent external radiation, and a few cases may even become resectable. Demonstration of the survival advantage will, however, require larger clinical trials with adequate controls. Some reports have suggested an unacceptably high rate of fatal hemoptysis following HDR endobronchial brachytherapy. Review of the world literature suggests that fatal hemoptysis rates range from 0-50% with an average of about 8%, comparable to an average of 5% with low dose rate brachytherapy. Other recognized complications include fistulae and radiation bronchitis. Because the majority of patients with

  7. Dose rate constants for new dose quantities

    International Nuclear Information System (INIS)

    Tschurlovits, M.; Daverda, G.; Leitner, A.

    1992-01-01

    Conceptual changes and new quantities made is necessary to reassess dose rate quantities. Calculations of the dose rate constant were done for air kerma, ambient dose equivalent and directional dose equivalent. The number of radionuclides is more than 200. The threshold energy is selected as 20 keV for the dose equivalent constants. The dose rate constant for the photon equivalent dose as used mainly in German speaking countries as a temporary quantity is also included. (Author)

  8. Effect of dose rate on residual γ-H2AX levels and frequency of micronuclei in X-irradiated mouse lymphocytes.

    Science.gov (United States)

    Turner, H C; Shuryak, I; Taveras, M; Bertucci, A; Perrier, J R; Chen, C; Elliston, C D; Johnson, G W; Smilenov, L B; Amundson, S A; Brenner, D J

    2015-03-01

    The biological risks associated with low-dose-rate (LDR) radiation exposures are not yet well defined. To assess the risk related to DNA damage, we compared the yields of two established biodosimetry end points, γ-H2AX and micronuclei (MNi), in peripheral mouse blood lymphocytes after prolonged in vivo exposure to LDR X rays (0.31 cGy/min) vs. acute high-dose-rate (HDR) exposure (1.03 Gy/min). C57BL/6 mice were total-body irradiated with 320 kVP X rays with doses of 0, 1.1, 2.2 and 4.45 Gy. Residual levels of total γ-H2AX fluorescence in lymphocytes isolated 24 h after the start of irradiation were assessed using indirect immunofluorescence methods. The terminal deoxynucleotidyl transferase dUTP nick end labeling (TUNEL) assay was used to determine apoptotic cell frequency in lymphocytes sampled at 24 h. Curve fitting analysis suggested that the dose response for γ-H2AX yields after acute exposures could be described by a linear dependence. In contrast, a linear-quadratic dose-response shape was more appropriate for LDR exposure (perhaps reflecting differences in repair time after different LDR doses). Dose-rate sparing effects (P effect across the dose range 24 h or 7 days post exposure. In conclusion, the γ-H2AX biomarker showed higher sensitivity to measure dose-rate effects after low-dose LDR X rays compared to MNi formation; however, confounding factors such as variable repair times post exposure, increased cell killing and cell cycle block likely contributed to the yields of MNi with accumulating doses of ionizing radiation.

  9. Commissioning and clinical implementation of HDR brachytherapy in El Salvador

    International Nuclear Information System (INIS)

    Morales Lopez, Jorge Luis; Castillo, Luis Frank; Castillo Bahi, Ramon del

    2009-01-01

    The Gynecologic Cancer is one of the best known malignancies in different countries of the world, with a high incidence in developing countries. In the treatment of this disease have been used multiple treatment arms among which is the high rate brachytherapy (HDR). The IAEA has put much emphasis on supporting all programs to treat this disease and in this context within the project 'Human Resource Development and Nuclear Technology Support', collaborated with the dispatch of experts on mission ELS0006 01 'Assistance to the ICES in HDR brachytherapy initiating Treatments at the Cancer Institute of El Salvador 'Dr. Narciso Diaz Bazan' in San Salvador, El Salvador. The process of commissioning and implementing clinical service Brachytherapy High Dose Rate (HDR BT) is a relatively complex process that begins with the formation of functional and technical service, based on flow patients to be treated, availability of local technological capability to install, and culminates with the preparation and implementation of protocols. Experts involved in the implementation of this service divided this task in stages organized chronologically: 1st. Study of existing infrastructure and level of training of technical personnel available, 2nd. Proposal and application of amendments in order to adapt the facility to the planned patient flow and optimal use of technological infrastructure, 3rd. Establishment of the process of securing the disposable waste materials and not required, 4th. Performance of tests of physical commissioning clinical dosimetry and instrumentation unit, surgical and therapeutic, 5th. Training of technical personnel, 6th. Preparation of clinical protocol and 7th. Initiation and development of treatment for patients. All these steps are carried out with the integration and consensus of the entire multidisciplinary team that makes up the service and with the support of the administration as a prerequisite. Within two weeks the service was modified according to

  10. Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source

    Energy Technology Data Exchange (ETDEWEB)

    Buchauer, Konrad; Henke, Guido; Schiefer, Hans; Plasswilm, Ludwig [Kantonsspital St. Gallen, Department of Radiation Oncology, St. Gallen (Switzerland)

    2014-12-15

    Conical surface applicators with an Ir-192 high-dose-rate brachytherapy source are a common modality for the treatment of non-melanomatous skin cancer with high tumour control rates. Surface dose characterisation of the Varian Varisource GammaMed+ IX afterloader vertical type surface applicators is performed two dimensionally using high-resolution film dosimetry. The focus of this study was to determine if Varian surface applicators with a vertical source suffer from the dose distribution irregularities reported for comparable applicators. Our goal was to evaluate if the irregularities found affected treatment and dose output verification procedures. Ionisation chamber-based verification of applicator output was established according to guidelines provided by the manufacturer. For additional measurement of surface dose Gafchromic EBT3 film dosimetry was used. The term ''therapeutic dose'' was defined as 85 % of the prescribed dose level. For the 10 different applicator inserts evaluated, cold spots were observed. Mean cold spot size was 2.0 mm x 3.6 mm (± 0.6 mm). The cold spots were dosimetrically well below 85 % of the prescribed dose. The cold spot was situated 2.2 mm (1.4-2.7 mm) unilaterally from the central axis and caused general asymmetry in the dose profiles intersecting the cold spot area. A source tilt of approximately 8 (± 1 ) was determined for the source used for irradiation. A central underdosed area exceeding 15 % of the prescribed dose has not been previously reported. Source tilt was observed and found to affect clinical use and possibly treatment outcome in applicators using a vertically arranged source. Surface applicators with a vertically orientated source were subject to dose irregularities that could impact on chamber-based applicator output verification procedures. We recommend film dosimetry-backed applicator commissioning to avoid systematic errors. (orig.) [German] Konische Oberflaechenapplikatoren sind ein

  11. Comparison of air-kerma strength determinations for HDR 192Ir sources

    International Nuclear Information System (INIS)

    Rasmussen, Brian E.; Davis, Stephen D.; Schmidt, Cal R.; Micka, John A.; DeWerd, Larry A.

    2011-01-01

    Purpose: To perform a comparison of the interim air-kerma strength standard for high dose rate (HDR) 192 Ir brachytherapy sources maintained by University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL) with measurements of the various source models using modified techniques from the literature. The current interim standard was established by Goetsch et al. in 1991 and has remained unchanged to date. Methods: The improved, laser-aligned seven-distance apparatus of University of Wisconsin Medical Radiation Research Center (UWMRRC) was used to perform air-kerma strength measurements of five different HDR 192 Ir source models. The results of these measurements were compared with those from well chambers traceable to the original standard. Alternative methodologies for interpolating the 192 Ir air-kerma calibration coefficient from the NIST air-kerma standards at 137 Cs and 250 kVp x rays (M250) were investigated and intercompared. As part of the interpolation method comparison, the Monte Carlo code EGSnrc was used to calculate updated values of A wall for the Exradin A3 chamber used for air-kerma strength measurements. The effects of air attenuation and scatter, room scatter, as well as the solution method were investigated in detail. Results: The average measurements when using the inverse N K interpolation method for the Classic Nucletron, Nucletron microSelectron, VariSource VS2000, GammaMed Plus, and Flexisource were found to be 0.47%, -0.10%, -1.13%, -0.20%, and 0.89% different than the existing standard, respectively. A further investigation of the differences observed between the sources was performed using MCNP5 Monte Carlo simulations of each source model inside a full model of an HDR 1000 Plus well chamber. Conclusions: Although the differences between the source models were found to be statistically significant, the equally weighted average difference between the seven-distance measurements and the well chambers was 0.01%, confirming that

  12. Comparison of air-kerma strength determinations for HDR (192)Ir sources.

    Science.gov (United States)

    Rasmussen, Brian E; Davis, Stephen D; Schmidt, Cal R; Micka, John A; Dewerd, Larry A

    2011-12-01

    To perform a comparison of the interim air-kerma strength standard for high dose rate (HDR) (192)Ir brachytherapy sources maintained by the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL) with measurements of the various source models using modified techniques from the literature. The current interim standard was established by Goetsch et al. in 1991 and has remained unchanged to date. The improved, laser-aligned seven-distance apparatus of the University of Wisconsin Medical Radiation Research Center (UWMRRC) was used to perform air-kerma strength measurements of five different HDR (192)Ir source models. The results of these measurements were compared with those from well chambers traceable to the original standard. Alternative methodologies for interpolating the (192)Ir air-kerma calibration coefficient from the NIST air-kerma standards at (137)Cs and 250 kVp x rays (M250) were investigated and intercompared. As part of the interpolation method comparison, the Monte Carlo code EGSnrc was used to calculate updated values of A(wall) for the Exradin A3 chamber used for air-kerma strength measurements. The effects of air attenuation and scatter, room scatter, as well as the solution method were investigated in detail. The average measurements when using the inverse N(K) interpolation method for the Classic Nucletron, Nucletron microSelectron, VariSource VS2000, GammaMed Plus, and Flexisource were found to be 0.47%, -0.10%, -1.13%, -0.20%, and 0.89% different than the existing standard, respectively. A further investigation of the differences observed between the sources was performed using MCNP5 Monte Carlo simulations of each source model inside a full model of an HDR 1000 Plus well chamber. Although the differences between the source models were found to be statistically significant, the equally weighted average difference between the seven-distance measurements and the well chambers was 0.01%, confirming that it is not necessary to

  13. Viewpoint adaptive display of HDR images

    DEFF Research Database (Denmark)

    Forchhammer, Søren; Mantel, Claire

    2017-01-01

    In this paper viewpoint adaptive display of HDR images incorporating the effects of ambient light is presented and evaluated. LED backlight displays may render HDR images, but while at a global scale a high dynamic range may be achieved, locally the contrast is limited by the leakage of light...... through the LC elements of the display. To render high quality images, the display with backlight dimming can compute the values of the LED backlight and LC elements based on the input image, information about the viewpoint of the observer(s) and information of the ambient light. The goal is to achieve...... the best perceptual reproduction of the specified target image derived from the HDR input image in the specific viewing situation including multiple viewers, possibly having different preferences. An optimization based approach is presented. Some tests with reproduced images are also evaluated subjectively...

  14. Optimum radiotherapy schedule for uterine cervical cancer based-on the detailed information of dose fractionation and radiotherapy technique

    International Nuclear Information System (INIS)

    Cho, Jae Ho; Kim, Hyun Chang; Suh, Chang Ok

    2005-01-01

    The best dose-fractionation regimen of the definitive radiotherapy for cervix cancer remains to be clearly determined. It seems to be partially attributed to the complexity of the affecting factors and the lack of detailed information on external and intra-cavitary fractionation. To find optimal practice guidelines, our experiences of the combination of external beam radiotherapy (EBRT) and high-dose-rate intracavitary brachytherapy (HDR-ICBT) were reviewed with detailed information of the various treatment parameters obtained from a large cohort of women treated homogeneously at a single institute. The subjects were 743 cervical cancer patients (Stage IB 198, IIA 77, IIB 364, IIIA 7, IIIB 89 and IVA 8) treated by radiotherapy alone, between 1990 and 1996. A total external beam radiotherapy (EBRT) dose of 23.4 ∼ 59.4 Gy (Median 45.0) was delivered to the whole pelvis. High-dose-rate intracavitary brachytherapy (HDR-ICBT) was also performed using various fractionation schemes. A Midline block (MLB) was initiated after the delivery of 14.4∼ 43.2 Gy (Median 36.0) of EBRT in 495 patients, while in the other 248 patients EBRT could not be used due to slow tumor regression or the huge initial bulk of tumor. The point A, actual bladder and rectal doses were individually assessed in all patients. The biologically effective dose (BED) to the tumor (α / β = 10) and late-responding tissues (α /β = 3) for both EBRT and HDR-ICBT were calculated. The total BED values to point A, the actual bladder and rectal reference points were the summation of the EBRT and HDR-ICBT. In addition to all the details on dose-fractionation, the other factors (i.e. the overall treatment time, physicians preference) that can affect the schedule of the definitive radiotherapy were also thoroughly analyzed. The association between MD-BED Gy 3 and the risk of complication was assessed using serial multiple logistic regressions models. The associations between R-BED Gy 3 and rectal complications

  15. Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources.

    Science.gov (United States)

    Kumar, Sudhir; Datta, D; Sharma, S D; Chourasiya, G; Babu, D A R; Sharma, D N

    2014-04-01

    Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of sensitive volume 0.6 cm(3) is one of the recommended methods for measuring RAKR of HDR (192)Ir brachytherapy sources. While using the cylindrical chamber method, it is required to determine the positioning error of the ionization chamber with respect to the source which is called the distance error. An attempt has been made to apply the fuzzy set theory to estimate the subjective uncertainty associated with the distance error. A simplified approach of applying this fuzzy set theory has been proposed in the quantification of uncertainty associated with the distance error. In order to express the uncertainty in the framework of fuzzy sets, the uncertainty index was estimated and was found to be within 2.5%, which further indicates that the possibility of error in measuring such distance may be of this order. It is observed that the relative distance li estimated by analytical method and fuzzy set theoretic approach are consistent with each other. The crisp values of li estimated using analytical method lie within the bounds computed using fuzzy set theory. This indicates that li values estimated using analytical methods are within 2.5% uncertainty. This value of uncertainty in distance measurement should be incorporated in the uncertainty budget, while estimating the expanded uncertainty in HDR (192)Ir source strength measurement.

  16. HDR video synthesis for vision systems in dynamic scenes

    Science.gov (United States)

    Shopovska, Ivana; Jovanov, Ljubomir; Goossens, Bart; Philips, Wilfried

    2016-09-01

    High dynamic range (HDR) image generation from a number of differently exposed low dynamic range (LDR) images has been extensively explored in the past few decades, and as a result of these efforts a large number of HDR synthesis methods have been proposed. Since HDR images are synthesized by combining well-exposed regions of the input images, one of the main challenges is dealing with camera or object motion. In this paper we propose a method for the synthesis of HDR video from a single camera using multiple, differently exposed video frames, with circularly alternating exposure times. One of the potential applications of the system is in driver assistance systems and autonomous vehicles, involving significant camera and object movement, non- uniform and temporally varying illumination, and the requirement of real-time performance. To achieve these goals simultaneously, we propose a HDR synthesis approach based on weighted averaging of aligned radiance maps. The computational complexity of high-quality optical flow methods for motion compensation is still pro- hibitively high for real-time applications. Instead, we rely on more efficient global projective transformations to solve camera movement, while moving objects are detected by thresholding the differences between the trans- formed and brightness adapted images in the set. To attain temporal consistency of the camera motion in the consecutive HDR frames, the parameters of the perspective transformation are stabilized over time by means of computationally efficient temporal filtering. We evaluated our results on several reference HDR videos, on synthetic scenes, and using 14-bit raw images taken with a standard camera.

  17. A fast inverse treatment planning strategy facilitating optimized catheter selection in image-guided high-dose-rate interstitial gynecologic brachytherapy.

    Science.gov (United States)

    Guthier, Christian V; Damato, Antonio L; Hesser, Juergen W; Viswanathan, Akila N; Cormack, Robert A

    2017-12-01

    Interstitial high-dose rate (HDR) brachytherapy is an important therapeutic strategy for the treatment of locally advanced gynecologic (GYN) cancers. The outcome of this therapy is determined by the quality of dose distribution achieved. This paper focuses on a novel yet simple heuristic for catheter selection for GYN HDR brachytherapy and their comparison against state of the art optimization strategies. The proposed technique is intended to act as a decision-supporting tool to select a favorable needle configuration. The presented heuristic for catheter optimization is based on a shrinkage-type algorithm (SACO). It is compared against state of the art planning in a retrospective study of 20 patients who previously received image-guided interstitial HDR brachytherapy using a Syed Neblett template. From those plans, template orientation and position are estimated via a rigid registration of the template with the actual catheter trajectories. All potential straight trajectories intersecting the contoured clinical target volume (CTV) are considered for catheter optimization. Retrospectively generated plans and clinical plans are compared with respect to dosimetric performance and optimization time. All plans were generated with one single run of the optimizer lasting 0.6-97.4 s. Compared to manual optimization, SACO yields a statistically significant (P ≤ 0.05) improved target coverage while at the same time fulfilling all dosimetric constraints for organs at risk (OARs). Comparing inverse planning strategies, dosimetric evaluation for SACO and "hybrid inverse planning and optimization" (HIPO), as gold standard, shows no statistically significant difference (P > 0.05). However, SACO provides the potential to reduce the number of used catheters without compromising plan quality. The proposed heuristic for needle selection provides fast catheter selection with optimization times suited for intraoperative treatment planning. Compared to manual optimization, the

  18. High dose rate interstitial brachytherapy with external beam irradiation for localized prostate cancer. Preliminary results

    Energy Technology Data Exchange (ETDEWEB)

    Hiratsuka, Junichi; Jo, Yoshimasa; Yoden, Eisaku; Tanaka, Hiroyoshi; Imajo, Yoshinari [Kawasaki Medical School, Kurashiki, Okayama (Japan); Nagase, Naomi; Narihiro, Naomasa; Kubota, Juichi

    2000-12-01

    This study was undertaken to assess the biochemical and pathological results of combined external beam radiotherapy and high dose rate Ir-192 brachytherapy (HDR-Ir192) for clinically localized prostate cancer. Between October 1997 and August 1999, 39 evaluable patients with adenocarcinoma of prostate diagnosed by biopsy were treated with interstitial and external beam irradiation. Patients ranged in age from 58-82 years, with a mean of 69.7 years. T1c, T2 and T3 tumors, according to the UICC classification system (1997), were found in 7, 21 and 11 cases respectively. The mean initial pre-treatment PSA was 35.9 ng/ml (median 13.2), with 77% of the patients having had a pre-treatment PSA greater than 10 ng/ml. Of all patients, 17 had received pre-treatment hormonal therapy. Hormonal pretreatment was stopped at the beginning of radiotherapy in all cases. External beam four-field box irradiation was given to the small pelvis to a dose of 45 Gy/25 fractions. Three HDR-Ir192 treatments were given over a 30-h period, with 5.5 Gy per fraction at the circumference of the prostate gland over the course of this study. Biochemical failure was defined as a PSA level >1.5 ng/ml and rising on three consecutive values. If serial post-treatment PSA levels showed a continuous downward trend, failure was not scored. The patient with clinical evidence of progression was classified as a clinical failure. The median follow-up at the time of evaluation was 19.6 months. A post-treatment PSA level {<=}1.0 ng/ml was seen in 26 (67%) patients, and values from >1.0 to {<=}2.0 ng/ml were seen in 10 (26%) patients. Biochemical failure was not seen in 38 patients except for one patient who developed a distant bone metastasis with negative prostatic biopsy 15 months after treatment. Biochemical control rate was 100% (38/38) except for the patient with bone metastasis classified as clinical failure. Negative biopsies 18 months after treatment were found in 93% (14/15) of patients. Only one patient

  19. SU-F-T-37: Dosimetric Evaluation of Planned Versus Decay Corrected Treatment Plans for the Treatment of Tandem-Based Cervical HDR Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Goyal, M [Texas Oncology, PA, Fort Worth, TX (United States); Shobhit University, Meerut, Uttar Pradesh (India); Manjhi, J; Rai, D [Shobhit University, Meerut, Uttar Pradesh (India); Kehwar, T [Pinnacle Health Cancer Center, Mechanicsburg, PA (United States); Barker, J; Heintz, B; Shide, K [Texas Oncology, PA, Fort Worth, TX (United States)

    2016-06-15

    Purpose: This study evaluated dosimetric parameters for actual treatment plans versus decay corrected treatment plans for cervical HDR brachytherapy. Methods: 125 plans of 25 patients, who received 5 fractions of HDR brachytherapy, were evaluated in this study. Dose was prescribed to point A (ICRU-38) and High risk clinical tumor volume (HR-CTV) and organs at risk (OAR) were, retrospectively, delineated on original CT images by treating physician. First HDR plan was considered as reference plan and decay correction was applied to calculate treatment time for subsequent fractions, and was applied, retrospectively, to determine point A, HR-CTV D90, and rectum and bladder doses. Results: The differences between mean point A reference doses and the point A doses of the plans computed using decay times were found to be 1.05%±0.74% (−2.26% to 3.26%) for second fraction; −0.25%±0.84% (−3.03% to 3.29%) for third fraction; 0.04%±0.70% (−2.68% to 2.56%) for fourth fraction and 0.30%±0.81% (−3.93% to 2.67%) for fifth fraction. Overall mean point A dose difference, for all fractions, was 0.29%±0.38% (within ± 5%). Mean rectum and bladder dose differences were calculated to be −3.46%±0.12% and −2.47%±0.09%, for points, respectively, and −1.72%±0.09% and −0.96%±0.06%, for D2cc, respectively. HR-CTV D90 mean dose difference was found to be −1.67% ± 0.11%. There was no statistically significant difference between the reference planned point A doses and that calculated using decay time to the subsequent fractions (p<0.05). Conclusion: This study reveals that a decay corrected treatment will provide comparable dosimetric results and can be utilized for subsequent fractions of cervical HDR brachytherapy instead of actual treatment planning. This approach will increase efficiency, decrease workload, reduce patient observation time between applicator insertion and treatment delivery. This would be particularly useful for institutions with limited

  20. Changes in heparin dose response slope during cardiac surgery: possible result in inaccuracy in predicting heparin bolus dose requirement to achieve target ACT.

    Science.gov (United States)

    Ichikawa, Junko; Mori, Tetsu; Kodaka, Mitsuharu; Nishiyama, Keiko; Ozaki, Makoto; Komori, Makiko

    2017-09-01

    The substantial interpatient variability in heparin requirement has led to the use of a heparin dose response (HDR) technique. The accuracy of Hepcon-based heparin administration in achieving a target activated clotting time (ACT) using an HDR slope remains controversial. We prospectively studied 86 adult patients scheduled for cardiac surgery requiring cardiopulmonary bypass. The total dose of calculated heparin required for patient and pump priming was administered simultaneously to achieve a target ACT of 450 s for HDR on the Hepcon HMS system. Blood samples were obtained after the induction of anesthesia, at 3 min after heparin administration and after the initiation of CPB to measure kaolin ACT, HDR slope, whole-blood heparin concentration based on the HDR slope and anti-Xa heparin concentration, antithrombin and complete blood count. The target ACT of 450 s was not achieved in 68.6% of patients. Compared with patients who achieved the target ACT, those who failed to achieve their target ACT had a significantly higher platelet count at baseline. Correlation between the HDR slope and heparin sensitivity was poor. Projected heparin concentration and anti-Xa heparin concentration are not interchangeable based on the Bland-Altman analysis. It can be hypothesized that the wide discrepancy in HDR slope versus heparin sensitivity may be explained by an inaccurate prediction of the plasma heparin level and/or the change in HDR of individual patients, depending on in vivo factors such as extravascular sequestration of heparin, decreased intrinsic antithrombin activity level and platelet count and/or activity.

  1. High-Dose-Rate Brachytherapy Boost Effect on Local Tumor Control in Young Women With Breast Cancer

    International Nuclear Information System (INIS)

    Guinot, Jose-Luis; Baixauli-Perez, Cristobal; Soler, Pablo; Tortajada, Maria Isabel; Moreno, Araceli; Santos, Miguel Angel; Mut, Alejandro; Gozalbo, Francisco; Arribas, Leoncio

    2015-01-01

    Purpose: To evaluate the local control rate and complications of a single fraction of high-dose-rate brachytherapy (HDR BT) boost in women aged 45 yeas and younger after breast-conserving therapy. Methods and Materials: Between 1999 and 2007, 167 patients between the ages of 26 and 45 years old (72 were 40 years old or younger), with stages T1 to T2 invasive breast cancer with disease-free margin status of at least 5 mm after breast-conserving surgery received 46 to 50 Gy whole-breast irradiation plus a 7-Gy HDR-BT boost (“fast boostâ€). An axillary dissection was performed in 72.5% of the patients and sentinel lymph node biopsy in 27.5%. A supraclavicular area was irradiated in 19% of the patients. Chemotherapy was used in 86% of the patients and hormone treatment in 77%. Clinical nodes were present in 18% and pathological nodes in 29%. The pathological stage was pT0: 5%, pTis: 3%, pT1: 69% and pT2: 23%. Intraductal component was present in 40% and 28% were G3. Results: At a median follow-up of 92 months, 9 patients relapsed on the margin of the implant, and 1 patient in another quadrant, resulting in a 10-year local relapse rate of 4.3% and a breast relapse rate of 4.9%, with breast preservation in 93.4%; no case of mastectomy due to poor cosmesis arose. Actuarial 5- and 10-year disease-free, cause-specific, and overall survival rates were 87.9% and 85.8%, and 92.1% and 88.4%, and 92.1% and 87.3%, respectively. In a univariate analysis, triple-negative cases and negative hormone receptors did worse, but in a multivariate analysis, only the last factor was significant for local and breast control. Asymptomatic fibrosis G2 was recorded in 3 cases, and there were no other late complications. Cosmetic results were good to excellent in 97% of cases. Conclusions: A single dose of 7 Gy using the fast-boost technique is well tolerated, with a low rate of late complications and improved local tumor control in women aged 45 and younger, compared to published data

  2. High-Dose-Rate Brachytherapy Boost Effect on Local Tumor Control in Young Women With Breast Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Guinot, Jose-Luis, E-mail: jguinot@fivo.org [Department of Radiation Oncology, Fundacion Instituto Valenciano de Oncologia, Valencia (Spain); Baixauli-Perez, Cristobal [Health Services Research Unit, Center for Public Health Research, Valencia (Spain); Soler, Pablo; Tortajada, Maria Isabel; Moreno, Araceli; Santos, Miguel Angel; Mut, Alejandro [Department of Radiation Oncology, Fundacion Instituto Valenciano de Oncologia, Valencia (Spain); Gozalbo, Francisco [Department of Pathology, Fundacion Instituto Valenciano de Oncologia, Valencia (Spain); Arribas, Leoncio [Department of Radiation Oncology, Fundacion Instituto Valenciano de Oncologia, Valencia (Spain)

    2015-01-01

    Purpose: To evaluate the local control rate and complications of a single fraction of high-dose-rate brachytherapy (HDR BT) boost in women aged 45 yeas and younger after breast-conserving therapy. Methods and Materials: Between 1999 and 2007, 167 patients between the ages of 26 and 45 years old (72 were 40 years old or younger), with stages T1 to T2 invasive breast cancer with disease-free margin status of at least 5 mm after breast-conserving surgery received 46 to 50 Gy whole-breast irradiation plus a 7-Gy HDR-BT boost (“fast boostâ€). An axillary dissection was performed in 72.5% of the patients and sentinel lymph node biopsy in 27.5%. A supraclavicular area was irradiated in 19% of the patients. Chemotherapy was used in 86% of the patients and hormone treatment in 77%. Clinical nodes were present in 18% and pathological nodes in 29%. The pathological stage was pT0: 5%, pTis: 3%, pT1: 69% and pT2: 23%. Intraductal component was present in 40% and 28% were G3. Results: At a median follow-up of 92 months, 9 patients relapsed on the margin of the implant, and 1 patient in another quadrant, resulting in a 10-year local relapse rate of 4.3% and a breast relapse rate of 4.9%, with breast preservation in 93.4%; no case of mastectomy due to poor cosmesis arose. Actuarial 5- and 10-year disease-free, cause-specific, and overall survival rates were 87.9% and 85.8%, and 92.1% and 88.4%, and 92.1% and 87.3%, respectively. In a univariate analysis, triple-negative cases and negative hormone receptors did worse, but in a multivariate analysis, only the last factor was significant for local and breast control. Asymptomatic fibrosis G2 was recorded in 3 cases, and there were no other late complications. Cosmetic results were good to excellent in 97% of cases. Conclusions: A single dose of 7 Gy using the fast-boost technique is well tolerated, with a low rate of late complications and improved local tumor control in women aged 45 and younger, compared to published data

  3. Optimal bladder filling during high-dose-rate intracavitary brachytherapy for cervical cancer: a dosimetric study

    Directory of Open Access Journals (Sweden)

    Umesh Mahantshetty

    2017-04-01

    Full Text Available Purpose: The aim of this study is to compare 3D dose volume histogram (DVH parameters of bladder and other organs at risk with different bladder filling protocol during high-dose-rate intracavitary brachytherapy (HDR-ICBT in cervical cancer, and to find optimized bladder volume. Material and methods : This dosimetric study was completed with 21 patients who underwent HDR-ICBT with computed tomography/magnetic resonance compatible applicator as a routine treatment. Computed tomography planning was done for each patient with bladder emptied (series 1, after 50 ml (series 2, and 100 ml (series 3 bladder filling with a saline infusion through the bladder catheter. Contouring was done on the Eclipse Planning System. 7 Gy to point A was prescribed with the standard loading patterns. Various 3D DVH parameters including 0.1 cc, 1 cc, 2 cc doses and mean doses to the OAR’s were noted. Paired t-test was performed. Results : The mean (± SD bladder volume was 64.5 (± 25 cc, 116.2 (± 28 cc, and 172.9 (± 29 cc, for series 1, 2, and 3, respectively. The 0.1 cm 3 ,1 cm 3 , 2 cm 3 mean bladder doses for series 1, series 2, and series 3 were 9.28 ± 2.27 Gy, 7.38 ± 1.72 Gy, 6.58 ± 1.58 Gy; 9.39 ± 2.28 Gy, 7.85 ± 1.85 Gy, 7.05 ± 1.59 Gy, and 10.09 ± 2.46 Gy, 8.33 ± 1.75 Gy, 7.6 ± 1.55 Gy, respectively. However, there was a trend towards higher bladder doses in series 3. Similarly, for small bowel dose 0.1 cm 3 , 1 cm 3 , and 2 cm 3 in series 1, 2, and 3 were 5.44 ± 2.2 Gy, 4.41 ± 1.84 Gy, 4 ± 1.69 Gy; 4.57 ± 2.89 Gy, 3.78 ± 2.21 Gy, 3.35 ± 2.02 Gy, and 4.09 ± 2.38 Gy, 3.26 ± 1.8 Gy, 3.05 ± 1.58 Gy. Significant increase in small bowel dose in empty bladder (series 1 compared to full bladder (series 3 (p = 0.03 was noted. However, the rectal and sigmoid doses were not significantly affected with either series. Conclusions : Bladder filling protocol with 50 ml and 100 ml was well tolerated and achieved a reasonably reproducible bladder volume

  4. Performance evaluation of objective quality metrics for HDR image compression

    Science.gov (United States)

    Valenzise, Giuseppe; De Simone, Francesca; Lauga, Paul; Dufaux, Frederic

    2014-09-01

    Due to the much larger luminance and contrast characteristics of high dynamic range (HDR) images, well-known objective quality metrics, widely used for the assessment of low dynamic range (LDR) content, cannot be directly applied to HDR images in order to predict their perceptual fidelity. To overcome this limitation, advanced fidelity metrics, such as the HDR-VDP, have been proposed to accurately predict visually significant differences. However, their complex calibration may make them difficult to use in practice. A simpler approach consists in computing arithmetic or structural fidelity metrics, such as PSNR and SSIM, on perceptually encoded luminance values but the performance of quality prediction in this case has not been clearly studied. In this paper, we aim at providing a better comprehension of the limits and the potentialities of this approach, by means of a subjective study. We compare the performance of HDR-VDP to that of PSNR and SSIM computed on perceptually encoded luminance values, when considering compressed HDR images. Our results show that these simpler metrics can be effectively employed to assess image fidelity for applications such as HDR image compression.

  5. SU-E-T-525: Dose Volume Histograms (DVH) Analysis and Comparison with ICRU Point Doses in MRI Guided HDR Brachytherapy for Cervical Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Badkul, R; McClinton, C; Kumar, P; Mitchell, M [University of Kansas Medical Center, Kansas City, KS (United States)

    2014-06-01

    .1 overestimated the bladder ICRU point dose up to 43% for conventional-techniques.Bladder-D2 provided a good estimation of ICRU bladder point-doses(within 3.6%) for conventional-techniques. This correlation is not observed for MGO plans perhaps due to steering of isodose line, leading to unpredictable dwell-weighting. Conclusion: MRI based HDR-planning provides accurate delineation of tumor volumes and normal structures, and optimized tumor-coverage can be achieved with acceptable normal-tissue doses. This study showed that for conventional techniques D0.1 rectum dose and D2 bladder dose are good representation of ICRU-reference-point doses.

  6. Dose rate correction in medium dose rate brachytherapy for carcinoma cervix

    International Nuclear Information System (INIS)

    Patel, F.D.; Negi, P.S.; Sharma, S.C.; Kapoor, R.; Singh, D.P.; Ghoshal, S.

    1998-01-01

    Purpose: To establish the magnitude of brachytherapy dose reduction required for stage IIB and III carcinoma cervix patients treated by external radiation and medium dose rate (MDR) brachytherapy at a dose rate of 220±10 cGy/h at point A.Materials and methods: In study-I, at the time of MDR brachytherapy application at a dose rate of 220±10 cGy/h at point A, patients received either 3060 cGy, a 12.5% dose reduction (MDR-12.5), or 2450 cGy, a 30% dose reduction (MDR-30), to point A and they were compared to a group of previously treated LDR patients who received 3500 cGy to point A at a dose rate of 55-65 cGy/h. Study-II was a prospective randomized trial and patients received either 2450 cGy, a 30% dose reduction (MDR-II (30)) or 2800 cGy, a 20% dose reduction (MDR-II (20)), at point A. Patients were evaluated for local control of disease and morbidity. Results: In study-I the 5-year actuarial local control rate in the MDR-30 and MDR-12.5 groups was 71.7±10% and 70.5±10%, respectively, compared to 63.4±10% in the LDR group. However, the actuarial morbidity (all grades) in the MDR-12.5 group was 58.5±14% as against 34.9±9% in the LDR group (P 3 developed complication as against 62.5% of those receiving a rectal BED of (140 3 (χ 2 =46.43; P<0.001). Conclusion: We suggest that at a dose rate of 220±10 cGy/h at point A the brachytherapy dose reduction factor should be around 30%, as suggested by radiobiological data, to keep the morbidity as low as possible without compromising the local control rates. (Copyright (c) 1998 Elsevier Science B.V., Amsterdam. All rights reserved.)

  7. Variations of intracavitary applicator geometry during multiple HDR brachytherapy insertions in carcinoma cervix and its influence on reporting as per ICRU report 38

    International Nuclear Information System (INIS)

    Datta, Niloy Ranjan; Kumar, Shaleen; Das, Koilpillai Joseph Maria; Pandey, Chandra Mani; Halder, Shikha; Ayyagari, Sunder

    2001-01-01

    Purpose: This paper examines the extent of variation in the applicator geometry during multiple high dose rate (HDR) intracavitary brachytherapy (ICBT) applications and its impact on reporting as per ICRU report 38. Materials and methods: Eighty orthogonal radiographs from 20 consecutive patients of carcinoma cervix (FIGO stages, IIA-IIIB) having four HDR ICBT applications of 6 Gy each at weekly intervals following teletherapy were evaluated. The applicator consisted of a flexible intrauterine tandem (IUT) independent of the ovoid assembly. The applicator geometry was evaluated in terms of: α angle, β angle, intrauterine length (IUTL), interovoid (IOV), os to right ovoid (ORT) and os to left ovoid (OLT) distances along with vertical (VDL) and anteroposterior displacements (ADL) of the os with respect to the ovoids. The Cartesian co-ordinates (X, Y, and Z) of the IUT tip, centre of both ovoids and os were also measured. Doses to right point A (ARD), left point A (ALD), along with a reference volume of 6 Gy for ICRU height (IRH), width (IRW), thickness (IRT) and volume (IRV) were estimated for each application. Results: Highly significant differences (P<0.001) between four insertions in any given patient across 20 patients for α angle, β angle, IUTL, IOV, ORT, VDL, co-ordinates of the IUT, ovoids and os were observed, except for ADL (P=0.041) and OLT (P=0.247). As a consequence, variations were observed in ARD (P=0.027), ALD (P=0.017); IRH, IRW, IRT and IRV (all P<0.001). Applicator factors which influenced the various dose specification parameters were: β angle and ORT for both ARD and ALD; UTLN, VDL and ORT for IRH; UTLN and IOV for IRW; UTLN for IRT and VDL for the 6 Gy IRV. Conclusions: A significant variation of the applicator geometry and its movement was observed in patients undergoing multiple HDR ICBT. This could have implications for reporting dose and volume specifications as required by ICRU report 38

  8. Comparison of air-kerma strength determinations for HDR {sup 192}Ir sources

    Energy Technology Data Exchange (ETDEWEB)

    Rasmussen, Brian E.; Davis, Stephen D.; Schmidt, Cal R.; Micka, John A.; DeWerd, Larry A. [Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706 (United States)

    2011-12-15

    Purpose: To perform a comparison of the interim air-kerma strength standard for high dose rate (HDR) {sup 192}Ir brachytherapy sources maintained by University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL) with measurements of the various source models using modified techniques from the literature. The current interim standard was established by Goetsch et al. in 1991 and has remained unchanged to date. Methods: The improved, laser-aligned seven-distance apparatus of University of Wisconsin Medical Radiation Research Center (UWMRRC) was used to perform air-kerma strength measurements of five different HDR {sup 192}Ir source models. The results of these measurements were compared with those from well chambers traceable to the original standard. Alternative methodologies for interpolating the {sup 192}Ir air-kerma calibration coefficient from the NIST air-kerma standards at {sup 137}Cs and 250 kVp x rays (M250) were investigated and intercompared. As part of the interpolation method comparison, the Monte Carlo code EGSnrc was used to calculate updated values of A{sub wall} for the Exradin A3 chamber used for air-kerma strength measurements. The effects of air attenuation and scatter, room scatter, as well as the solution method were investigated in detail. Results: The average measurements when using the inverse N{sub K} interpolation method for the Classic Nucletron, Nucletron microSelectron, VariSource VS2000, GammaMed Plus, and Flexisource were found to be 0.47%, -0.10%, -1.13%, -0.20%, and 0.89% different than the existing standard, respectively. A further investigation of the differences observed between the sources was performed using MCNP5 Monte Carlo simulations of each source model inside a full model of an HDR 1000 Plus well chamber. Conclusions: Although the differences between the source models were found to be statistically significant, the equally weighted average difference between the seven-distance measurements and the well

  9. Dosimetric measurements of an 192Ir HDR source with a diamond detector

    International Nuclear Information System (INIS)

    Rustgi, Surendra N.

    1996-01-01

    Purpose: To study the feasibility of using a diamond detector for the dosimetry of a high dose rate (HDR) 192 Ir source and to compare the measurement results with published data and calculations from a commercial treatment planning system. Materials and methods: The sensitive volume of the diamond detector consists of a disk of 0.26 mm thickness and 3 mm diameter. The detector was applied an external bias of +100 V and was preirradiated to a dose of 500 cGy to stabilize its response. The 192 Ir source from the Nucletron microSelectron unit has an active diameter of 0.6 mm and a length of 3.5 mm. Photon fluence anisotropy factors in air were measured at distances of 5 and 10 cm from two sources and compared with TLD measurements. Dose profiles and isodose distributions were measured at several distances from the source and compared with calculations from a Nucletron treatment planning system. These dose calculations in water use a point source approximation with the anisotropy factors independent of the radial distance from the source. Results: The photon fluence around the 192 Ir HDR source, measured with a diamond detector at distances of 5 and 10 cm from the source, is very anisotropic. Compared to the source transverse direction, the photon fluence intensity along the source axis reduces to approximately 60%. Measurements performed on two sources indicate that the photon anisotropy does not change with distance in air. Within experimental uncertainty, similar results were obtained with TLD rods and are in excellent agreement with published anisotropy factors 1 . Dose profiles, measured with the diamond detector in a water phantom, at distances of 1,2,3 and 5 cm from the source, are found to be in excellent agreement with the Nucletron planning system calculations. Similar excellent agreement is observed between the measured and calculated isodose curves in planes parallel to the source plane. Conclusion: The diamond detector has been demonstrated to be suitable

  10. HDR Image Quality Enhancement Based on Spatially Variant Retinal Response

    Directory of Open Access Journals (Sweden)

    Horiuchi Takahiko

    2010-01-01

    Full Text Available There is a growing demand for being able to display high dynamic range (HDR images on low dynamic range (LDR devices. Tone mapping is a process for enhancing HDR image quality on an LDR device by converting the tonal values of the original image from HDR to LDR. This paper proposes a new tone mapping algorithm for enhancing image quality by deriving a spatially-variant operator for imitating S-potential response in human retina, which efficiently improves local contrasts while conserving good global appearance. The proposed tone mapping operator is studied from a system construction point of view. It is found that the operator is regarded as a natural extension of the Retinex algorithm by adding a global adaptation process to the local adaptation. The feasibility of the proposed algorithm is examined in detail on experiments using standard HDR images and real HDR scene images, comparing with conventional tone mapping algorithms.

  11. Development of a program for calculation of second dose and securities in brachytherapy high dose rate

    International Nuclear Information System (INIS)

    Esteve Sanchez, S.; Martinez Albaladejo, M.; Garcia Fuentes, J. D.; Bejar Navarro, M. J.; Capuz Suarez, B.; Moris de Pablos, R.; Colmenares Fernandez, R.

    2015-01-01

    We assessed the reliability of the program with 80 patients in the usual points of prescription of each pathology. The average error of the calculation points is less than 0.3% in 95% of cases, finding the major differences in the axes of the applicators (maximum error -0.798%). The program has proved effective previously testing him with erroneous dosimetry. Thanks to the implementation of this program is achieved by the calculation of the dose and part of the process of quality assurance program in a few minutes, highlighting the case of HDR prostate due to having a limited time. Having separate data sheet allows each institution to its protocols modify parameters. (Author)

  12. Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source

    Directory of Open Access Journals (Sweden)

    Barlanka Ravikumar

    2012-01-01

    Full Text Available In Brachytherapy treatment planning, the effects of tissue heterogeneities are commonly neglected due to lack of accurate, general and fast three-dimensional (3D dose-computational algorithms. In performing dose calculations, it is assumed that the tumor and surrounding tissues constitute a uniform, homogeneous medium equivalent to water. In the recent past, three-dimensional computed tomography (3D-CT based treatment planning for Brachytherapy applications has been popularly adopted. However, most of the current commercially available planning systems do not provide the heterogeneity corrections for Brachytherapy dosimetry. In the present study, we have measured and quantified the impact of inhomogeneity caused by different tissues with a 0.015 cc ion chamber. Measurements were carried out in wax phantom which was employed to measure the heterogeneity. Iridium-192 (192 Ir source from high dose rate (HDR Brachytherapy machine was used as the radiation source. The reduction of dose due to tissue inhomogeneity was measured as the ratio of dose measured with different types of inhomogeneity (bone, spleen, liver, muscle and lung to dose measured with homogeneous medium for different distances. It was observed that different tissues attenuate differently, with bone tissue showing maximum attenuation value and lung tissue resulting minimum value and rest of the tissues giving values lying in between those of bone and lung. It was also found that inhomogeneity at short distance is considerably more than that at larger distances.

  13. Proposal of a high dose rate brachytherapy model for in vitro radiobiology studies

    International Nuclear Information System (INIS)

    Geraldo, Jony M.; Nogueira, Luciana B.; Andrade, Lidia M.; Trindade, Cassia; Furtado, Clascidia A.; Ladeira, Luiz Orlando

    2016-01-01

    The aim of this research was to develop an easy and reproducible approach for experimental HDR brachytherapy allowing in vitro irradiation studies based on clinical parameters. An acrylic platform was designed to attach T25 tissue culture flasks and multi-well tissue culture plates as well as kept the catheters in a fixed position during irradiation. CT images were taken and the irradiation was planned for 550cGy dose applied on adherent tumor cells. Dosimetric measurements were done and all relevant uncertainties were taken into account in order to figure out the correct dose range received by the cells. Tumor cells were irradiated two times over an interval of 24h between irradiations. Proof of concepts of this approach was carried out by biological effects analysis using a radioresistant human epidermoid carcinoma A431 cell line. Cellular proliferation and cell cycle phase were assessed by Trypan blue exclusion assay and DNA content analysis by flow cytometry, respectively. This approach allowed uniform dose distribution around the arrangement in all types of tissue culture plastics evaluated. Corrections due to uncertainties were managed. Regarding in vitro assays there was a significant (p<0.05) decreasing of cellular proliferation rate in irradiated cells. Moreover, increased percentage of cells arrested in G2/M phase (32.3 ± 1.5%) were observed for treated group compared with untreated cells. (author)

  14. Development of a TLD mailed system for remote dosimetry audit for 192Ir HDR and PDR sources

    International Nuclear Information System (INIS)

    Roue, Amelie; Venselaar, Jack L.M.; Ferreira, Ivaldo H.; Bridier, Andre; Dam, Jan van

    2007-01-01

    Background and purpose: In the framework of an ESTRO ESQUIRE project, the BRAPHYQS Physics Network and the EQUAL-ESTRO laboratory have developed a procedure for checking the absorbed dose to water in the vicinity of HDR or PDR sources using a mailed TLD system. The methodology and the materials used in the procedure are based on the existing EQUAL-ESTRO external radiotherapy dose checks. Materials and methods: A phantom for TLD postal dose assurance service, adapted to accept catheters from different HDR afterloaders, has been developed. The phantom consists of three PMMA tubes supporting catheters placed at 120 degrees around a central TLD holder. A study on the use of LiF powder type DTL 937 (Philitech) has been performed in order to establish the TLD calibration in dose-to-water at a given distance from 192 Ir source, as well as to determine all correction factors to convert the TLD reading into absorbed dose to water. The dosimetric audit is based on the comparison between the dose to water measured with the TL dosimeter and the dose calculated by the clinical TPS. Results of the audits are classified in four different levels depending on the ratio of the measured dose to the stated dose. The total uncertainty budget in the measurement of the absorbed dose to water using TLD near an 192 Ir HDR source, including TLD reading, correction factors and TLD calibration coefficient, is determined as 3.27% (1 s). Results: To validate the procedures, the external audit was first tested among the members of the BRAPHYQS Network. Since November 2004, the test has been made available for use by all European brachytherapy centres. To date, 11 centres have participated in the checks and the results obtained are very encouraging. Nevertheless, one error detected has shown the usefulness of this audit. Conclusion: A method of absorbed dose to water determination in the vicinity of an 192 Ir brachytherapy source was developed for the purpose of a mailed TL dosimetry system. The

  15. Development of a TLD mailed system for remote dosimetry audit for (192)Ir HDR and PDR sources.

    Science.gov (United States)

    Roué, Amélie; Venselaar, Jack L M; Ferreira, Ivaldo H; Bridier, André; Van Dam, Jan

    2007-04-01

    In the framework of an ESTRO ESQUIRE project, the BRAPHYQS Physics Network and the EQUAL-ESTRO laboratory have developed a procedure for checking the absorbed dose to water in the vicinity of HDR or PDR sources using a mailed TLD system. The methodology and the materials used in the procedure are based on the existing EQUAL-ESTRO external radiotherapy dose checks. A phantom for TLD postal dose assurance service, adapted to accept catheters from different HDR afterloaders, has been developed. The phantom consists of three PMMA tubes supporting catheters placed at 120 degrees around a central TLD holder. A study on the use of LiF powder type DTL 937 (Philitech) has been performed in order to establish the TLD calibration in dose-to-water at a given distance from (192)Ir source, as well as to determine all correction factors to convert the TLD reading into absorbed dose to water. The dosimetric audit is based on the comparison between the dose to water measured with the TL dosimeter and the dose calculated by the clinical TPS. Results of the audits are classified in four different levels depending on the ratio of the measured dose to the stated dose. The total uncertainty budget in the measurement of the absorbed dose to water using TLD near an (192)Ir HDR source, including TLD reading, correction factors and TLD calibration coefficient, is determined as 3.27% (1s). To validate the procedures, the external audit was first tested among the members of the BRAPHYQS Network. Since November 2004, the test has been made available for use by all European brachytherapy centres. To date, 11 centres have participated in the checks and the results obtained are very encouraging. Nevertheless, one error detected has shown the usefulness of this audit. A method of absorbed dose to water determination in the vicinity of an (192)Ir brachytherapy source was developed for the purpose of a mailed TL dosimetry system. The accuracy of the procedure was determined. This method allows a

  16. A comparative study of quality of life in patients with localized prostate cancer treated at a single institution: Stereotactic ablative radiotherapy or external beam + high dose rate brachytherapy boost

    International Nuclear Information System (INIS)

    Helou, Joelle; Morton, Gerard; Zhang, Liying; Deabreu, Andrea; D’Alimonte, Laura; Elias, Evelyn; Musunuru, Hima Bindu; Mamedov, Alexandre; Ravi, Ananth; Chung, Hans; Cheung, Patrick; Loblaw, Andrew

    2014-01-01

    Purpose: To compare the quality of life (QOL) in patients treated with stereotactic ablative radiation therapy (SABR) alone or high dose rate (HDR) brachytherapy + hypofractionated external beam radiotherapy (EBRT). Methods and materials: Patient self-reported QOL was prospectively measured among patients from two sequential phase 2 clinical trials: 1-SABR 35 Gy/5 fractions/5 weeks, 2–15 Gy HDR 1 fraction, followed by EBRT 37.5 Gy/15 fractions/3 weeks. The expanded prostate cancer index composite was assessed at baseline and q6 monthly up to 5 years. Urinary, bowel and sexual domains were analyzed. A minimally clinical important change (MCIC) was defined as 0.5*standard deviation of the baseline for each domain. Fisher exact test and general linear mixed model were used (p < 0.05). Results: 84 and 123 patients were treated on the SABR and HDR boost studies, with a median follow up of 51 and 61 months respectively. There was a significant difference in MCIC between treatments in the urinary function and bother (p < 0.0001), the bowel function (p = 0.0216) and the sexual function (p = 0.0419) and bother (p = 0.0290) domains in favor of the SABR group. Of patients who reported no problem with their sexual function at baseline, 7% and 23% respectively considered it to be a moderate to big problem on follow up (p = 0.0077). Conclusion: Patients treated with HDR-boost reported deterioration of QOL particularly in sexual domains in comparison with SABR

  17. WE-F-BRD-01: HDR Brachytherapy II: Integrating Imaging with HDR

    International Nuclear Information System (INIS)

    Craciunescu, O; Todor, D; Leeuw, A de

    2014-01-01

    In recent years, with the advent of high/pulsed dose rate afterloading technology, advanced treatment planning systems, CT/MRI compatible applicators, and advanced imaging platforms, image-guided adaptive brachytherapy treatments (IGABT) have started to play an ever increasing role in modern radiation therapy. The most accurate way to approach IGABT treatment is to provide the infrastructure that combines in a single setting an appropriate imaging device, a treatment planning system, and a treatment unit. The Brachytherapy Suite is not a new concept, yet the modern suites are incorporating state-of-the-art imaging (MRI, CBCT equipped simulators, CT, and /or US) that require correct integration with each other and with the treatment planning and delivery systems. Arguably, an MRI-equipped Brachytherapy Suite is the ideal setup for real-time adaptive brachytherapy treatments. The main impediment to MRI-IGABT adoption is access to MRI scanners. Very few radiation oncology departments currently house MRI scanners, and even fewer in a dedicated Brachytherapy Suite. CBCT equipped simulators are increasingly offered by manufacturers as part of a Brachytherapy Suite installation. If optimized, images acquired can be used for treatment planning, or can be registered with other imaging modalities. This infrastructure is relevant for all forms of brachytherapy, especially those utilizing multi-fractionated courses of treatment such as prostate and cervix. Moreover, for prostate brachytherapy, US imaging systems can be part of the suite to allow for real-time HDR/LDR treatments. Learning Objectives: Understand the adaptive workflow of MR-based IGBT for cervical cancer. Familiarize with commissioning aspects of a CBCT equipped simulator with emphasis on brachytherapy applications Learn about the current status and future developments in US-based prostate brachytherapy

  18. WE-F-BRD-01: HDR Brachytherapy II: Integrating Imaging with HDR

    Energy Technology Data Exchange (ETDEWEB)

    Craciunescu, O [Duke University Medical Center, Durham, NC (United States); Todor, D [Virginia Commonwealth University, Richmond, VA (United States); Leeuw, A de

    2014-06-15

    In recent years, with the advent of high/pulsed dose rate afterloading technology, advanced treatment planning systems, CT/MRI compatible applicators, and advanced imaging platforms, image-guided adaptive brachytherapy treatments (IGABT) have started to play an ever increasing role in modern radiation therapy. The most accurate way to approach IGABT treatment is to provide the infrastructure that combines in a single setting an appropriate imaging device, a treatment planning system, and a treatment unit. The Brachytherapy Suite is not a new concept, yet the modern suites are incorporating state-of-the-art imaging (MRI, CBCT equipped simulators, CT, and /or US) that require correct integration with each other and with the treatment planning and delivery systems. Arguably, an MRI-equipped Brachytherapy Suite is the ideal setup for real-time adaptive brachytherapy treatments. The main impediment to MRI-IGABT adoption is access to MRI scanners. Very few radiation oncology departments currently house MRI scanners, and even fewer in a dedicated Brachytherapy Suite. CBCT equipped simulators are increasingly offered by manufacturers as part of a Brachytherapy Suite installation. If optimized, images acquired can be used for treatment planning, or can be registered with other imaging modalities. This infrastructure is relevant for all forms of brachytherapy, especially those utilizing multi-fractionated courses of treatment such as prostate and cervix. Moreover, for prostate brachytherapy, US imaging systems can be part of the suite to allow for real-time HDR/LDR treatments. Learning Objectives: Understand the adaptive workflow of MR-based IGBT for cervical cancer. Familiarize with commissioning aspects of a CBCT equipped simulator with emphasis on brachytherapy applications Learn about the current status and future developments in US-based prostate brachytherapy.

  19. Dosimetric analysis at ICRU reference points in HDR-brachytherapy of cervical carcinoma.

    Science.gov (United States)

    Eich, H T; Haverkamp, U; Micke, O; Prott, F J; Müller, R P

    2000-01-01

    In vivo dosimetry in bladder and rectum as well as determining doses on suggested reference points following the ICRU report 38 contribute to quality assurance in HDR-brachytherapy of cervical carcinoma, especially to minimize side effects. In order to gain information regarding the radiation exposure at ICRU reference points in rectum, bladder, ureter and regional lymph nodes those were calculated (digitalisation) by means of orthogonal radiographs of 11 applications in patients with cervical carcinoma, who received primary radiotherapy. In addition, the doses at the ICRU rectum reference point was compared to the results of in vivo measurements in the rectum. The in vivo measurements were by factor 1.5 below the doses determined for the ICRU rectum reference point (4.05 +/- 0.68 Gy versus 6.11 +/- 1.63 Gy). Reasons for this were: calibration errors, non-orthogonal radiographs, movement of applicator and probe in the time span between X-ray and application, missing connection of probe and anterior rectal wall. The standard deviation of calculations at ICRU reference points was on average +/- 30%. Possible reasons for the relatively large standard deviation were difficulties in defining the points, identifying them on radiographs and the different locations of the applicators. Although 3 D CT, US or MR based treatment planning using dose volume histogram analysis is more and more established, this simple procedure of marking and digitising the ICRU reference points lengthened treatment planning only by 5 to 10 minutes. The advantages of in vivo dosimetry are easy practicability and the possibility to determine rectum doses during radiation. The advantages of computer-aided planning at ICRU reference points are that calculations are available before radiation and that they can still be taken into account for treatment planning. Both methods should be applied in HDR-brachytherapy of cervical carcinoma.

  20. Quantitative analysis of biological responses to low dose-rate γ-radiation, including dose, irradiation time, and dose-rate

    International Nuclear Information System (INIS)

    Magae, J.; Furukawa, C.; Kawakami, Y.; Hoshi, Y.; Ogata, H.

    2003-01-01

    Full text: Because biological responses to radiation are complex processes dependent on irradiation time as well as total dose, it is necessary to include dose, dose-rate and irradiation time simultaneously to predict the risk of low dose-rate irradiation. In this study, we analyzed quantitative relationship among dose, irradiation time and dose-rate, using chromosomal breakage and proliferation inhibition of human cells. For evaluation of chromosome breakage we assessed micronuclei induced by radiation. U2OS cells, a human osteosarcoma cell line, were exposed to gamma-ray in irradiation room bearing 50,000 Ci 60 Co. After the irradiation, they were cultured for 24 h in the presence of cytochalasin B to block cytokinesis, cytoplasm and nucleus were stained with DAPI and propidium iodide, and the number of binuclear cells bearing micronuclei was determined by fluorescent microscopy. For proliferation inhibition, cells were cultured for 48 h after the irradiation and [3H] thymidine was pulsed for 4 h before harvesting. Dose-rate in the irradiation room was measured with photoluminescence dosimeter. While irradiation time less than 24 h did not affect dose-response curves for both biological responses, they were remarkably attenuated as exposure time increased to more than 7 days. These biological responses were dependent on dose-rate rather than dose when cells were irradiated for 30 days. Moreover, percentage of micronucleus-forming cells cultured continuously for more than 60 days at the constant dose-rate, was gradually decreased in spite of the total dose accumulation. These results suggest that biological responses at low dose-rate, are remarkably affected by exposure time, that they are dependent on dose-rate rather than total dose in the case of long-term irradiation, and that cells are getting resistant to radiation after the continuous irradiation for 2 months. It is necessary to include effect of irradiation time and dose-rate sufficiently to evaluate risk

  1. ELDRS Characterization for a Very High Dose Mission

    Science.gov (United States)

    Harris, Richard D.; McClure, Steven S.; Rax, Bernard G.; Kenna, Aaron J.; Thorbourn, Dennis O.; Clark, Karla B.; Yan, Tsun-Yee

    2010-01-01

    Evaluation of bipolar linear parts which may have Enhanced Low Dose Rate Sensitivity (ELDRS) is problematic for missions that have very high dose radiation requirements. The accepted standards for evaluating parts that display ELDRS require testing at a very low dose rate which could be prohibitively long for very high dose missions. In this work, a methodology for ELDRS characterization of bipolar parts for mission doses up to 1 Mrad(Si) is evaluated. The procedure employs an initial dose rate of 0.01 rad(Si)/s to a total dose of 50 krad(Si) and then changes to 0.04 rad(Si)/s to a total dose of 1 Mrad(Si). This procedure appears to work well. No change in rate of degradation with dose has been observed when the dose rate is changed from 0.01 to 0.04 rad(Si)/s. This is taken as an indication that the degradation due to the higher dose rate is equivalent to that at the lower dose rate at the higher dose levels, at least for the parts studied to date. In several cases, significant parameter degradation or functional failure not observed at HDR was observed at fairly high total doses (50 to 250 krad(Si)) at LDR. This behavior calls into question the use of dose rate trend data and enhancement factors to predict LDR performance.

  2. Sci-Thur AM: YIS – 03: Combining sagittally-reconstructed 3D and live-2D ultrasound for high-dose-rate prostate brachytherapy needle segmentation

    Energy Technology Data Exchange (ETDEWEB)

    Hrinivich, Thomas; Hoover, Douglas; Surry, Kathleen; Edirisinghe, Chandima; D’Souza, David; Fenster, Aaron; Wong, Eugene [University of Western Ontario, London Regional Cancer Program/LHSC, London Regional Cancer Program/LHSC, Robarts Research Institute, London Regional Cancer Program/LHSC, Robarts Research Institute, University of Western Ontario (Canada)

    2016-08-15

    Ultrasound-guided high-dose-rate prostate brachytherapy (HDR-BT) needle segmentation is performed clinically using live-2D sagittal images. Organ segmentation is then performed using axial images, introducing a source of geometric uncertainty. Sagittally-reconstructed 3D (SR3D) ultrasound enables both needle and organ segmentation, but suffers from shadow artifacts. We present a needle segmentation technique augmenting SR3D with live-2D sagittal images using mechanical probe tracking to mitigate image artifacts and compare it to the clinical standard. Seven prostate cancer patients underwent TRUS-guided HDR-BT during which the clinical and proposed segmentation techniques were completed in parallel using dual ultrasound video outputs. Calibrated needle end-length measurements were used to calculate insertion depth errors (IDEs), and the dosimetric impact of IDEs was evaluated by perturbing clinical treatment plan source positions. The proposed technique provided smaller IDEs than the clinical approach, with mean±SD of −0.3±2.2 mm and −0.5±3.7mm respectively. The proposed and clinical techniques resulted in 84% and 43% of needles with IDEs within ±3mm, and IDE ranges across all needles of [−7.7mm, 5.9mm] and [−9.3mm, 7.7mm] respectively. The proposed and clinical IDEs lead to mean±SD changes in the volume of the prostate receiving the prescription dose of −0.6±0.9% and −2.0±5.3% respectively. The proposed technique provides improved HDR-BT needle segmentation accuracy over the clinical technique leading to decreased dosimetric uncertainty by eliminating the axial-to-sagittal registration, and mitigates the effect of shadow artifacts by incorporating mechanically registered live-2D sagittal images.

  3. Color sensitivity of the multi-exposure HDR imaging process

    Science.gov (United States)

    Lenseigne, Boris; Jacobs, Valéry Ann; Withouck, Martijn; Hanselaer, Peter; Jonker, Pieter P.

    2013-04-01

    Multi-exposure high dynamic range(HDR) imaging builds HDR radiance maps by stitching together different views of a same scene with varying exposures. Practically, this process involves converting raw sensor data into low dynamic range (LDR) images, estimate the camera response curves, and use them in order to recover the irradiance for every pixel. During the export, applying white balance settings and image stitching, which both have an influence on the color balance in the final image. In this paper, we use a calibrated quasi-monochromatic light source, an integrating sphere, and a spectrograph in order to evaluate and compare the average spectral response of the image sensor. We finally draw some conclusion about the color consistency of HDR imaging and the additional steps necessary to use multi-exposure HDR imaging as a tool to measure the physical quantities such as radiance and luminance.

  4. Monte Carlo characterization of the Gamma-Med Hdr plus Ir-192 brachytherapy source

    Energy Technology Data Exchange (ETDEWEB)

    Reyes, E.; Sosa, M. A.; Gil V, A. [Universidad de Guanajuato, Division de Ciencias e Ingenierias, Av. Insurgentes 2354, 37150 Leon, Guanajuato (Mexico); Monzon, E., E-mail: eric_1985@fisica.ugto.mx [IMSS, Unidad Medica de Alta Especialidad No. 1, Av. Adolfo Lopez Mateos 1813, 37340 Leon, Guanajuato (Mexico)

    2015-10-15

    Full text: The MCNP4C Monte Carlo code was used to simulate the dosimetry around the Gamma-Med Hdr Plus iridium-192 brachytherapy source in both air/vacuum and water environments. Dosimetry data in water was calculated and are presented into an away-along table. All dosimetric quantities recommended by the AAPM Task Group 43 report have been also calculated. These quantities are air kerma strength, dose rate constant, radial dose function and anisotropy function. The obtained data are compared to this source reference data, finding results in good agreement with them. In this study, recommendations of the AAPM TG-43U1 report have been followed and comply with the most recent AAPM and ESTRO physics committee recommendations about Monte Carlo techniques. The data in the present study complement published data and can be used as input in the Tps or as benchmark data to verify the results of the treatment planning systems as well as a means of comparison with other datasets from this source. (Author)

  5. Monte Carlo characterization of the Gamma-Med Hdr plus Ir-192 brachytherapy source

    International Nuclear Information System (INIS)

    Reyes, E.; Sosa, M. A.; Gil V, A.; Monzon, E.

    2015-10-01

    Full text: The MCNP4C Monte Carlo code was used to simulate the dosimetry around the Gamma-Med Hdr Plus iridium-192 brachytherapy source in both air/vacuum and water environments. Dosimetry data in water was calculated and are presented into an away-along table. All dosimetric quantities recommended by the AAPM Task Group 43 report have been also calculated. These quantities are air kerma strength, dose rate constant, radial dose function and anisotropy function. The obtained data are compared to this source reference data, finding results in good agreement with them. In this study, recommendations of the AAPM TG-43U1 report have been followed and comply with the most recent AAPM and ESTRO physics committee recommendations about Monte Carlo techniques. The data in the present study complement published data and can be used as input in the Tps or as benchmark data to verify the results of the treatment planning systems as well as a means of comparison with other datasets from this source. (Author)

  6. Radiobiological aspects of continuous low dose-rate irradiation and fractionated high dose-rate irradiation

    International Nuclear Information System (INIS)

    Turesson, I.

    1990-01-01

    The biological effects of continuous low dose-rate irradiation and fractionated high dose-rate irradiation in interstitial and intracavitary radiotherapy and total body irradiation are discussed in terms of dose-rate fractionation sensitivity for various tissues. A scaling between dose-rate and fraction size was established for acute and late normal-tissue effects which can serve as a guideline for local treatment in the range of dose rates between 0.02 and 0.005 Gy/min and fraction sizes between 8.5 and 2.5 Gy. This is valid provided cell-cycle progression and proliferation can be ignored. Assuming that the acute and late tissue responses are characterized by α/β values of about 10 and 3 Gy and a mono-exponential repair half-time of about 3 h, the same total doses given with either of the two methods are approximately equivalent. The equivalence for acute and late non-hemopoietic normal tissue damage is 0.02 Gy/min and 8.5 Gy per fraction; 0.01 Gy/min and 5.5 Gy per fraction; and 0.005 Gy/min and 2.5Gy per fraction. A very low dose rate, below 0.005 Gy/min, is thus necessary to simulate high dose-rate radiotherapy with fraction sizes of about 2Gy. The scaling factor is, however, dependent on the repair half-time of the tissue. A review of published data on dose-rate effects for normal tissue response showed a significantly stronger dose-rate dependence for late than for acute effects below 0.02 Gy/min. There was no significant difference in dose-rate dependence between various acute non-hemopoietic effects or between various late effects. The consistent dose-rate dependence, which justifies the use of a general scaling factor between fraction size and dose rate, contrasts with the wide range of values for repair half-time calculated for various normal-tissue effects. This indicates that the model currently used for repair kinetics is not satisfactory. There are also few experimental data in the clinical dose-rate range, below 0.02 Gy/min. It is therefore

  7. Commissioning of a well type chamber for HDR and LDR brachytherapy applications: a review of methodology and outcomes.

    Science.gov (United States)

    Mukwada, Godfrey; Neveri, Gabor; Alkhatib, Zaid; Waterhouse, David K; Ebert, Martin

    2016-03-01

    For safe and accurate dose delivery in brachytherapy, associated equipment is subject to commissioning and ongoing quality assurance (QA). Many centres depend on the use of a well-type chamber ('well chamber') for performing brachytherapy dosimetry. Documentation of well chamber commissioning is scarce despite the important role the chamber plays in the whole brachytherapy QA process. An extensive and structured commissioning of the HDR 1000 plus well chamber (Standard Imaging Inc, Middleton WI) for HDR and LDR dosimetry was undertaken at Sir Charles Gairdner Hospital. The methodology and outcomes of this commissioning is documented and presented as a guideline to others involved in brachytherapy. The commissioning tests described include mechanical integrity, leakage current, directional dependence, response, length of uniform response, the influence of insert holders, ion collection efficiency, polarity effect, accuracy of measured air kerma strength (S(K)) or reference air kerma rate (K(R)) and baseline setting (for ongoing constancy checks). For the HDR 1000 plus well chamber, some of the insert holders modify the response curve. The measured sweet length was 2.5 cm which is within 0.5% of that specified by the manufacturer. Correction for polarity was negligible (0.9999) and ion recombination was small (0.9994). Directional dependence was small (less than 0.2%) and leakage current was negligible. The measured K(R) for (192)Ir agreed within 0.11% compared with a second well chamber of similar model and was within 0.5% of that determined via a free-in-air measurement method. Routine constancy checks over a year agreed with the baseline within 0.4%.

  8. Accelerated partial breast irradiation: An analysis of variables associated with late toxicity and long-term cosmetic outcome after high-dose-rate interstitial brachytherapy

    International Nuclear Information System (INIS)

    Wazer, David E.; Kaufman, Seth; Cuttino, Laurie; Di Petrillo, Thomas; Arthur, Douglas W.

    2006-01-01

    Purpose: To perform a detailed analysis of variables associated with late tissue effects of high-dose-rate (HDR) interstitial brachytherapy accelerated partial breast irradiation (APBI) in a large cohort of patients with prolonged follow-up. Methods and Materials: Beginning in 1995, 75 women with Stage I/II breast cancer were enrolled in identical institutional trials evaluating APBI as monotherapy after lumpectomy. Patients eligible included those with T1-2, N0-1 (≤3 nodes positive), M0 tumors of nonlobular histology with negative surgical margins, no extracapsular nodal extension, and negative results on postexcision mammogram. All patients underwent surgical excision and postoperative irradiation with HDR interstitial brachytherapy. The planning target volume was defined as the excision cavity plus a 2-cm margin. Treatment was delivered with a high-activity Ir-192 source at 3.4 Gy per fraction twice daily for 5 days to a total dose of 34 Gy. Dosimetric analyses were performed with three-dimensional postimplant dose and volume reconstructions. All patients were evaluated at 3-6-month intervals and assessed with a standardized cosmetic rating scale and according to Radiation Therapy Oncology Group late normal tissue toxicity scoring criteria. Clinical and therapy-related features were analyzed for their relationship to cosmetic outcome and toxicity rating. Clinical features analyzed included age, volume of resection, history of diabetes or hypertension, extent of axillary surgery, and systemic therapies. Therapy-related features analyzed included volume of tissue encompassed by the 100%, 150%, and 200% isodose lines (V100, V150, and V200, respectively), the dose homogeneity index (DHI), number of source dwell positions, and planar separation. Results: The median follow-up of all patients was 73 months (range, 43-118 months). The cosmetic outcome at last follow-up was rated as excellent, good, and fair/poor in 67%, 24%, and 9% of patients, respectively

  9. Therapeutic analysis of high-dose-rate {sup 192}Ir vaginal cuff brachytherapy for endometrial cancer using a cylindrical target volume model and varied cancer cell distributions

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, Hualin, E-mail: hualin.zhang@northwestern.edu; Donnelly, Eric D.; Strauss, Jonathan B. [Department of Radiation Oncology, Robert H. Lurie Comprehensive Cancer Center, Northwestern University Feinberg School of Medicine, Northwestern Memorial Hospital, Chicago, Illinois 60611 (United States); Qi, Yujin [Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522 (Australia)

    2016-01-15

    Purpose: To evaluate high-dose-rate (HDR) vaginal cuff brachytherapy (VCBT) in the treatment of endometrial cancer in a cylindrical target volume with either a varied or a constant cancer cell distributions using the linear quadratic (LQ) model. Methods: A Monte Carlo (MC) technique was used to calculate the 3D dose distribution of HDR VCBT over a variety of cylinder diameters and treatment lengths. A treatment planning system (TPS) was used to make plans for the various cylinder diameters, treatment lengths, and prescriptions using the clinical protocol. The dwell times obtained from the TPS were fed into MC. The LQ model was used to evaluate the therapeutic outcome of two brachytherapy regimens prescribed either at 0.5 cm depth (5.5 Gy × 4 fractions) or at the vaginal mucosal surface (8.8 Gy × 4 fractions) for the treatment of endometrial cancer. An experimentally determined endometrial cancer cell distribution, which showed a varied and resembled a half-Gaussian distribution, was used in radiobiology modeling. The equivalent uniform dose (EUD) to cancer cells was calculated for each treatment scenario. The therapeutic ratio (TR) was defined by comparing VCBT with a uniform dose radiotherapy plan in term of normal cell survival at the same level of cancer cell killing. Calculations of clinical impact were run twice assuming two different types of cancer cell density distributions in the cylindrical target volume: (1) a half-Gaussian or (2) a uniform distribution. Results: EUDs were weakly dependent on cylinder size, treatment length, and the prescription depth, but strongly dependent on the cancer cell distribution. TRs were strongly dependent on the cylinder size, treatment length, types of the cancer cell distributions, and the sensitivity of normal tissue. With a half-Gaussian distribution of cancer cells which populated at the vaginal mucosa the most, the EUDs were between 6.9 Gy × 4 and 7.8 Gy × 4, the TRs were in the range from (5.0){sup 4} to (13

  10. Air kerma standard for calibration of well-type chambers in Brazil using {sup 192}Ir HDR sources and its traceability

    Energy Technology Data Exchange (ETDEWEB)

    Di Prinzio, Renato; Almeida, Carlos Eduardo de [Laboratorio de Ciencias Radiologicas-Universidade do Estado do Rio de Janeiro (LCR/UERJ), R. Sao Francisco Xavier, 524, Pavilhao Haroldo Lisboa da Cunha, Terreo, Sala 136-Maracana, CEP 20550-900-Rio de Janeiro/RJ-Rio de Janeiro, RJ (Brazil) and Instituto de Radioprotecao e Dosimetria-Comissao Nacional de Energia Nuclear (IRD/CNEN), Av. Salvador Allende, s/n, Jacarepagua-CE22780-160-Rio de Janeiro, RJ (Brazil); Laboratorio de Ciencias Radiologicas-Universidade do Estado do Rio de Janeiro (LCR/UERJ), R. Sao Francisco Xavier, 524, Pavilhao Haroldo Lisboa da Cunha, Terreo, Sala 136-Maracana, CEP 20550-900-Rio de Janeiro/RJ-Rio de Janeiro, RJ (Brazil)

    2009-03-15

    In Brazil there are over 100 high dose rate (HDR) brachytherapy facilities using well-type chambers for the determination of the air kerma rate of {sup 192}Ir sources. This paper presents the methodology developed and extensively tested by the Laboratorio de Ciencias Radiologicas (LCR) and presently in use to calibrate those types of chambers. The system was initially used to calibrate six well-type chambers of brachytherapy services, and the maximum deviation of only 1.0% was observed between the calibration coefficients obtained and the ones in the calibration certificate provided by the UWADCL. In addition to its traceability to the Brazilian National Standards, the whole system was taken to University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL) for a direct comparison and the same formalism to calculate the air kerma was used. The comparison results between the two laboratories show an agreement of 0.9% for the calibration coefficients. Three Brazilian well-type chambers were calibrated at the UWADCL, and by LCR, in Brazil, using the developed system and a clinical HDR machine. The results of the calibration of three well chambers have shown an agreement better than 1.0%. Uncertainty analyses involving the measurements made both at the UWADCL and LCR laboratories are discussed.

  11. Equivalence of hyperfractionated and continuous brachytherapy in a rat tumor model and remarkable effectiveness when preceded by external irradiation.

    NARCIS (Netherlands)

    Veninga, T.; Visser, A.G.; Berg, A.P. van den; Hooije, C.M. van; Geel, C.A. van; Levendag, P.C.

    2001-01-01

    PURPOSE: In clinical brachytherapy, there is a tendency to replace continuous low-dose-rate (LDR) irradiation by either single-dose or fractionated high-dose-rate (HDR) irradiation. In this study, the equivalence of LDR treatments and fractionated HDR (2 fractions/day) or pulsed-dose-rate (PDR, 4

  12. Advantages of high-dose rate (HDR) brachytherapy in treatment of prostate cancer

    Science.gov (United States)

    Molokov, A. A.; Vanina, E. A.; Tseluyko, S. S.

    2017-09-01

    One of the modern methods of preserving organs radiation treatment is brachytherapy. This article analyzes the results of prostate brachytherapy. These studies of the advantages of high dose brachytherapy lead to the conclusion that this method of radiation treatment for prostate cancer has a favorable advantage in comparison with remote sensing methods, and is competitive, preserving organs in comparison to surgical methods of treatment. The use of the method of polyfocal transperineal biopsy during the brachytherapy session provides information on the volumetric spread of prostate cancer and adjust the dosimetry plan taking into account the obtained data.

  13. SU-F-T-31: Shape and Isodose Distributions in Co60 HDR Brachytherapy for Different Utero-Vaginal Time Ratios

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez-Sprinberg, G [Faculty of Sciences, Montevideo, Montevideo (Venezuela, Bolivarian Republic of); Piriz, G [Hospital Pereyra Rossell, Montevideo, Montevideo (Venezuela, Bolivarian Republic of)

    2016-06-15

    Purpose: To optimize the dose in bladder and rectum and show the different shapes of the isodose volumes in Co60-HDR brachytherapy, considering different utero and vaginal sources dwell ratio times (TU:TV). Methods: Besides Ir192-HDR, new Co60-HDR sources are being incorporated. We considered different TU:TV times and computed the dosis in bladder, rectum and at the reference points of the Manchester system. Also, we calculated the isodose volume and shape in each case. We used a EZAG-BEBIG Co0.A86 model with TPS HDRplus3.0.4. and LCT42-7, LCT42-2(R,L) applicators. A reference dose RA= 1.00 Gy was given to the A-right point. We considered the TU:TV dwell time ratios 1:0.25, 1:0.33, 1:0.5, 1:1, 1:2, 1:3, and 1:4. Given TU:TV, the stop time at each dwell position is fixed for each applicator. Results: Increasing TU:TV systematically results in a decreasing of the dose in bladder and rectum, e.g. 9% and 27% reduction were found in 1:0.25 with respect to 1:1, while 12% and 34% increase were found in 1:4 with respect to 1:1. Also, the isodose volume parameters height (h), width (w), thickness (t) and volume (hwt) increased from the 1:0.25 case to the 1:4 value: hwt is 25% lower and 31% higher than the 1:1 reference volume in these cases. Also w decreased for higher TU:TV and may compromise the tumoral volume coverage, decreasing 17% in the 1:0.25 case compared to the 1:1 case. The shape of the isodose volume was obtained for the different TU:TV considered. Conclusion: We obtained the shape of isodose volumes for different TU:TV values in gynecological Co60-HDR. We studied the dose reduction in bladder and rectum for different TU:TV ratios. The volume parameters and hwt are strongly dependent on this ratio. This information is useful for a quantitative check of the TPS and as a starting point towards optimization.

  14. A Dose-Volume Analysis of Magnetic Resonance Imaging-Aided High-Dose-Rate Image-Based Interstitial Brachytherapy for Uterine Cervical Cancer

    International Nuclear Information System (INIS)

    Yoshida, Ken; Yamazaki, Hideya; Takenaka, Tadashi; Kotsuma, Tadayuki; Yoshida, Mineo; Furuya, Seiichi; Tanaka, Eiichi; Uegaki, Tadaaki; Kuriyama, Keiko; Matsumoto, Hisanobu; Yamada, Shigetoshi; Ban, Chiaki

    2010-01-01

    Purpose: To investigate the feasibility of our novel image-based high-dose-rate interstitial brachytherapy (HDR-ISBT) for uterine cervical cancer, we evaluated the dose-volume histogram (DVH) according to the recommendations of the Gynecological GEC-ESTRO Working Group for image-based intracavitary brachytherapy (ICBT). Methods and Materials: Between June 2005 and June 2007, 18 previously untreated cervical cancer patients were enrolled. We implanted magnetic resonance imaging (MRI)-available plastic applicators by our unique ambulatory technique. Total treatment doses were 30-36 Gy (6 Gy per fraction) combined with external beam radiotherapy (EBRT). Treatment plans were created based on planning computed tomography with MRI as a reference. DVHs of the high-risk clinical target volume (HR CTV), intermediate-risk CTV (IR CTV), and the bladder and rectum were calculated. Dose values were biologically normalized to equivalent doses in 2-Gy fractions (EQD 2 ). Results: The median D90 (HR CTV) and D90 (IR CTV) per fraction were 6.8 Gy (range, 5.5-7.5) and 5.4 Gy (range, 4.2-6.3), respectively. The median V100 (HR CTV) and V100 (IR CTV) were 98.4% (range, 83-100) and 81.8% (range, 64-93.8), respectively. When the dose of EBRT was added, the median D90 and D100 of HR CTV were 80.6 Gy (range, 65.5-96.6) and 62.4 Gy (range, 49-83.2). The D 2cc of the bladder was 62 Gy (range, 51.4-89) and of the rectum was 65.9 Gy (range, 48.9-76). Conclusions: Although the targets were advanced and difficult to treat effectively by ICBT, MRI-aided image-based ISBT showed favorable results for CTV and organs at risk compared with previously reported image-based ICBT results.

  15. A dose-volume analysis of magnetic resonance imaging-aided high-dose-rate image-based interstitial brachytherapy for uterine cervical cancer.

    Science.gov (United States)

    Yoshida, Ken; Yamazaki, Hideya; Takenaka, Tadashi; Kotsuma, Tadayuki; Yoshida, Mineo; Furuya, Seiichi; Tanaka, Eiichi; Uegaki, Tadaaki; Kuriyama, Keiko; Matsumoto, Hisanobu; Yamada, Shigetoshi; Ban, Chiaki

    2010-07-01

    To investigate the feasibility of our novel image-based high-dose-rate interstitial brachytherapy (HDR-ISBT) for uterine cervical cancer, we evaluated the dose-volume histogram (DVH) according to the recommendations of the Gynecological GEC-ESTRO Working Group for image-based intracavitary brachytherapy (ICBT). Between June 2005 and June 2007, 18 previously untreated cervical cancer patients were enrolled. We implanted magnetic resonance imaging (MRI)-available plastic applicators by our unique ambulatory technique. Total treatment doses were 30-36 Gy (6 Gy per fraction) combined with external beam radiotherapy (EBRT). Treatment plans were created based on planning computed tomography with MRI as a reference. DVHs of the high-risk clinical target volume (HR CTV), intermediate-risk CTV (IR CTV), and the bladder and rectum were calculated. Dose values were biologically normalized to equivalent doses in 2-Gy fractions (EQD(2)). The median D90 (HR CTV) and D90 (IR CTV) per fraction were 6.8 Gy (range, 5.5-7.5) and 5.4 Gy (range, 4.2-6.3), respectively. The median V100 (HR CTV) and V100 (IR CTV) were 98.4% (range, 83-100) and 81.8% (range, 64-93.8), respectively. When the dose of EBRT was added, the median D90 and D100 of HR CTV were 80.6 Gy (range, 65.5-96.6) and 62.4 Gy (range, 49-83.2). The D(2cc) of the bladder was 62 Gy (range, 51.4-89) and of the rectum was 65.9 Gy (range, 48.9-76). Although the targets were advanced and difficult to treat effectively by ICBT, MRI-aided image-based ISBT showed favorable results for CTV and organs at risk compared with previously reported image-based ICBT results. (c) 2010 Elsevier Inc. All rights reserved.

  16. The usefulness of metal markers for CTV-based dose prescription in high-dose-rate interstitial brachytherapy

    International Nuclear Information System (INIS)

    Yoshida, Ken; Mitomo, Masanori; Nose, Takayuki; Koizumi, Masahiko; Nishiyama, Kinji; Yoshida, Mineo

    2002-01-01

    We employ a clinical target volume (CTV)-based dose prescription for high-dose-rate (HDR) interstitial brachytherapy. However, it is not easy to define CTV and organs at risk (OAR) from X-ray film or CT scanning. To solve this problem, we have utilized metal markers since October 1999. Moreover, metal markers can help modify dose prescription. By regulating the doses to the metal markers, refining the dose prescription can easily be achieved. In this research, we investigated the usefulness of the metal markers. Between October 1999 and May 2001, 51 patients were implanted with metal markers at Osaka Medical Center for Cancer and Cardiovascular Diseases (OMCC), Osaka National Hospital (ONH) and Sanda City Hospital (SCH). Forty-nine patients (head and neck: 32; pelvis: 11; soft tissue: 3; breast: 3) using metal markers were analyzed. During operation, we implanted 179 metal markers (49 patients) to CTV and 151 markers (26 patients) to OAR. At treatment planning, CTV was reconstructed judging from the metal markers, applicator position and operation records. Generally, we prescribed the tumoricidal dose to an isodose surface that covers CTV. We also planned to limit the doses to OAR lower than certain levels. The maximum normal tissue doses were decided 80%, 150%, 100%, 50% and 200% of the prescribed doses for the rectum, the urethra, the mandible, the skin and the large vessel, respectively. The doses to the metal markers using CTV-based dose prescription were generated. These were compared with the doses theoretically calculated with the Paris system. Treatment results were also investigated. The doses to the 158 metal markers (42 patients) for CTV were higher than ''tumoricidal dose''. In 7 patients, as a result of compromised dose prescription, 9 markers were lower than the tumoricidal dose. The other 12 markers (7%) were excluded from dose evaluation because they were judged as miss-implanted. The doses to the 142 metal markers (24 patients) for OAR were lower

  17. Outcomes Associated With 3 Treatment Schedules of High-Dose-Rate Brachytherapy Monotherapy for Favorable-Risk Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Jawad, Maha Saada; Dilworth, Joshua T.; Gustafson, Gary S.; Ye, Hong; Wallace, Michelle [Department of Radiation Oncology, Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States); Martinez, Alvaro [Michigan HealthCare Professionals/21" s" t Century Oncology, Farmington Hills, Michigan (United States); Chen, Peter Y. [Department of Radiation Oncology, Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States); Krauss, Daniel J., E-mail: DKrauss@beaumont.edu [Department of Radiation Oncology, Oakland University William Beaumont School of Medicine, Royal Oak, Michigan (United States)

    2016-03-15

    Purpose: We report the outcomes associated with 3 high-dose-rate (HDR) brachytherapy regimens used as monotherapy for favorable-risk prostate cancer. Methods and Materials: Four hundred ninety-four patients with stage ≤T2b prostate cancer, Gleason score ≤7, and prostate-specific antigen levels ≤15 ng/mL underwent HDR brachytherapy as monotherapy. Of those, 319 received 38 Gy in 4 fractions, 79 received 24 Gy in 2 fractions, and 96 received 27 Gy in 2 fractions. Acute and chronic genitourinary (GU) and gastrointestinal (GI) toxicities were defined as side effects occurring ≤6 and >6 months, respectively, after radiation therapy (RT) and were graded according to the Common Terminology Criteria for Adverse Events version 3.0. The time to toxicity was calculated from the date of RT completion. Variables were analyzed with χ{sup 2} test. P values <.05 were considered significant. Results: The median overall follow-up time was 4 years (range, 5.5, 3.5, and 2.5 years for 38 Gy, 24 Gy, and 27 Gy, respectively, P<.001). Acute and chronic grade ≥2 GU and GI toxicity profiles were similar among groups. Acceptable rates of grade 2 GU toxicities were seen with overall acute/chronic frequency/urgency, dysuria, retention, incontinence, and hematuria rates of 14%/20%, 6%/7%, 7%/4%, 1.5%/2%, and 1.5%/7%, respectively. Minimal grade 3 and no grade 4 or 5 toxicities were seen. Grade 1, 2, and 3 chronic urethral stricture rates were 0.3%, 2%, and 1%, respectively. All GI toxicities were similar between groups, with overall rates of acute/chronic grade 2 diarrhea, rectal pain/tenesmus, rectal bleeding, and proctitis of 1%/1%, <1%/0.5%, 0%/2%, and <1%/1%, respectively. No grade 3, 4, or 5 toxicities were seen. All comparisons were similar for hormone-naïve patients. The median time to maximal GU/GI toxicity was similar between groups, ranging from 1 to 1.6 to 0.9 to 1.2 years, respectively. There were no differences in clinical outcomes between the 3 groups at 5

  18. Survival and Complication Rate of Radiation Therapy in Stage I and II Carcinoma of uterine Cervix

    International Nuclear Information System (INIS)

    Ma, Sun Young; Cho, Heung Lea; Sohn, Seung Chang

    1995-01-01

    Purpose : To analyze survival rate and late rectal and bladder complication for patients with stage with stage I and II carcinoma of uterine cervix treated by radiation alone or combined with chemotherapy. Materials and Methods : Between November 1984 and December 1993, 127 patients with stage I and II carcinoma of uterine cervix treated by radiation alone or combined therapy of radiation and chemotherapy. Retrospective analysis for survival rate was carried out on eligible 107 patients and review for complication was possible in 91 patients. The median follow-up was 47 months (range 3-118) and the median age of patients was 56 years (range 31-76). 26 patients were stage IB by FIGO classification. 40 were stage IIA and 41 were stage IIB. 86 cases were treated by radiation alone and 21 were treated by radiation and chemotherapy. 101 patients were treated with intracavitary radiation therapy (ICRT), of these, 80 were received low dose rate (LDR) ICRT and 21 were received high dose rate (HDR) ICRT. Of the patients who received LDR ICRT, 63 were treated by 1 intracavitary insertion and 17 were underwent 2 insertions. And we evaluated the external radiation dose and midline shield. Results : Acturial survival rate at 5 years was 92% for stage IB, 75% for stage IIA, 53% for stage IIB and 69% in all patients. Grade 1 rectal complications were developed in 20 cases(22%), grade 2 were in 22 cases (24%), 22 cases (24%) of grade 1 urinary complications and 17 cases (19%) of grade 2 urinary complications were observed But no patients had severe complications that needed surgical management or admission care. Maximum bladder dose for the group of patients with urinary complications was higher than that for the patients without urinary complications(7608cGy v 6960cGy, p<0.01). Maximum rectal dose for the group of patients with rectal complications was higher than that for the patients without urinary complications (7041cGy v 6269cGy, p<0.01). While there was no significant

  19. WE-DE-201-01: BEST IN PHYSICS (THERAPY): A Fast Multi-Target Inverse Treatment Planning Strategy Optimizing Dosimetric Measures for High-Dose-Rate (HDR) Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Guthier, C [Brigham and Women’s Hospital, Boston, MA (United States); University Medical Center Mannheim, Mannheim (Germany); Harvard Medical School, Boston, MA (United States); Damato, A; Viswanathan, A; Cormack, R [Dana Farber Cancer Institut/Brigham and Women’s Hospital, Boston, MA (United States); Harvard Medical School, Boston, MA (United States); Hesser, J [University Medical Center Mannheim, Mannheim (Germany)

    2016-06-15

    Purpose: Inverse treatment planning (ITP) for interstitial HDR brachytherapy of gynecologic cancers seeks to maximize coverage of the clinical target volumes (tumor and vagina) while respecting dose-volume-histogram related dosimetric measures (DMs) for organs at risk (OARs). Commercially available ITP tools do not support DM-based planning because it is computationally too expensive to solve. In this study we present a novel approach that allows fast ITP for gynecologic cancers based on DMs for the first time. Methods: This novel strategy is an optimization model based on a smooth DM-based objective function. The smooth approximation is achieved by utilizing a logistic function for the evaluation of DMs. The resulting nonconvex and constrained optimization problem is then optimized with a BFGS algorithm. The model was evaluated using the implant geometry extracted from 20 patient treatment plans under an IRB-approved retrospective study. For each plan, the final DMs were evaluated and compared to the original clinical plans. The CTVs were the contoured tumor volume and the contoured surface of the vagina. Statistical significance was evaluated with a one-sided paired Wilcoxon signed-rank test. Results: As did the clinical plans, all generated plans fulfilled the defined DMs for OARs. The proposed strategy showed a statistically significant improvement (p<0.001) in coverage of the tumor and vagina, with absolute improvements of related DMs of (6.9 +/− 7.9)% and (28.2 +/− 12.0)%, respectively. This was achieved with a statistically significant (p<0.01) decrease of the high-dose-related DM for the tumor. The runtime of the optimization was (2.3 +/− 2.0) seconds. Conclusion: We demonstrated using clinical data that our novel approach allows rapid DM-based optimization with improved coverage of CTVs with fewer hot spots. Being up to three orders of magnitude faster than the current clinical practice, the method dramatically shortens planning time.

  20. WE-DE-201-01: BEST IN PHYSICS (THERAPY): A Fast Multi-Target Inverse Treatment Planning Strategy Optimizing Dosimetric Measures for High-Dose-Rate (HDR) Brachytherapy

    International Nuclear Information System (INIS)

    Guthier, C; Damato, A; Viswanathan, A; Cormack, R; Hesser, J

    2016-01-01

    Purpose: Inverse treatment planning (ITP) for interstitial HDR brachytherapy of gynecologic cancers seeks to maximize coverage of the clinical target volumes (tumor and vagina) while respecting dose-volume-histogram related dosimetric measures (DMs) for organs at risk (OARs). Commercially available ITP tools do not support DM-based planning because it is computationally too expensive to solve. In this study we present a novel approach that allows fast ITP for gynecologic cancers based on DMs for the first time. Methods: This novel strategy is an optimization model based on a smooth DM-based objective function. The smooth approximation is achieved by utilizing a logistic function for the evaluation of DMs. The resulting nonconvex and constrained optimization problem is then optimized with a BFGS algorithm. The model was evaluated using the implant geometry extracted from 20 patient treatment plans under an IRB-approved retrospective study. For each plan, the final DMs were evaluated and compared to the original clinical plans. The CTVs were the contoured tumor volume and the contoured surface of the vagina. Statistical significance was evaluated with a one-sided paired Wilcoxon signed-rank test. Results: As did the clinical plans, all generated plans fulfilled the defined DMs for OARs. The proposed strategy showed a statistically significant improvement (p<0.001) in coverage of the tumor and vagina, with absolute improvements of related DMs of (6.9 +/− 7.9)% and (28.2 +/− 12.0)%, respectively. This was achieved with a statistically significant (p<0.01) decrease of the high-dose-related DM for the tumor. The runtime of the optimization was (2.3 +/− 2.0) seconds. Conclusion: We demonstrated using clinical data that our novel approach allows rapid DM-based optimization with improved coverage of CTVs with fewer hot spots. Being up to three orders of magnitude faster than the current clinical practice, the method dramatically shortens planning time.