WorldWideScience

Sample records for current measuring device

  1. Simulation of leakage current measurement on medical devices using helmholtz coil configuration with different current flow

    Science.gov (United States)

    Sutanto, E.; Chandra, F.; Dinata, R.

    2017-05-01

    Leakage current measurement which can follow IEC standard for medical device is one of many challenges to be answered. The IEC 60601-1 has defined that the limit for a leakage current for Medical Device can be as low as 10 µA and as high as 500 µA, depending on which type of contact (applied part) connected to the patient. Most people are using ELCB (Earth-leakage circuit breaker) for safety purpose as this is the most common and available safety device in market. One type of ELCB devices is RCD (Residual Current Device) and this RCD type can measure the leakage current directly. This work will show the possibility on how Helmholtz Coil Configuration can be made to be like the RCD. The possibility is explored by comparing the magnetic field formula from each device, then it proceeds with a simulation using software EJS (Easy Java Simulation). The simulation will make sure the concept of magnetic field current cancellation follows the RCD concept. Finally, the possibility of increasing the measurement’s sensitivity is also analyzed. The sensitivity is needed to see the possibility on reaching the minimum leakage current limit defined by IEC, 0.01mA.

  2. Simulation of leakage current measurement on medical devices using helmholtz coil configuration with different current flow

    International Nuclear Information System (INIS)

    Sutanto, E; Chandra, F; Dinata, R

    2017-01-01

    Leakage current measurement which can follow IEC standard for medical device is one of many challenges to be answered. The IEC 60601-1 has defined that the limit for a leakage current for Medical Device can be as low as 10 µA and as high as 500 µA, depending on which type of contact (applied part) connected to the patient. Most people are using ELCB (Earth-leakage circuit breaker) for safety purpose as this is the most common and available safety device in market. One type of ELCB devices is RCD (Residual Current Device) and this RCD type can measure the leakage current directly. This work will show the possibility on how Helmholtz Coil Configuration can be made to be like the RCD. The possibility is explored by comparing the magnetic field formula from each device, then it proceeds with a simulation using software EJS (Easy Java Simulation). The simulation will make sure the concept of magnetic field current cancellation follows the RCD concept. Finally, the possibility of increasing the measurement’s sensitivity is also analyzed. The sensitivity is needed to see the possibility on reaching the minimum leakage current limit defined by IEC, 0.01mA. (paper)

  3. Classification of methods for measuring current-voltage characteristics of semiconductor devices

    Directory of Open Access Journals (Sweden)

    Iermolenko Ia. O.

    2014-06-01

    Full Text Available It is shown that computer systems for measuring current-voltage characteristics are very important for semiconductor devices production. The main criteria of efficiency of such systems are defined. It is shown that efficiency of such systems significantly depends on the methods for measuring current-voltage characteristics of semiconductor devices. The aim of this work is to analyze existing methods for measuring current-voltage characteristics of semiconductor devices and to create the classification of these methods in order to specify the most effective solutions in terms of defined criteria. To achieve this aim, the most common classifications of methods for measuring current-voltage characteristics of semiconductor devices and their main disadvantages are considered. Automated and manual, continuous, pulse, mixed, isothermal and isodynamic methods for measuring current-voltage characteristics are analyzed. As a result of the analysis and generalization of existing methods the next classification criteria are defined: the level of automation, the form of measurement signals, the condition of semiconductor device during the measurements, and the use of mathematical processing of the measurement results. With the use of these criteria the classification scheme of methods for measuring current-voltage characteristics of semiconductor devices is composed and the most effective methods are specified.

  4. A cryogenic current-measuring device with nano-ampere resolution at the storage ring TARN II

    International Nuclear Information System (INIS)

    Tanabe, T.; Chida, K.; Shinada, K.

    1999-01-01

    In cooler-ring experiments, an accurate and non-destructive current measurement is essential for determining the reaction cross sections. The lowest current which can be measured by the DC current transformer commonly used so far is some μA. In order to measure a low-beam current from nA to μA, we made a cryogenic current-measuring device using a superconducting quantum interference devices (SQUID), and measured the circulating ion current at the cooler ring TARN II. This paper gives the design and performance of the device

  5. Passive fault current limiting device

    Science.gov (United States)

    Evans, Daniel J.; Cha, Yung S.

    1999-01-01

    A passive current limiting device and isolator is particularly adapted for use at high power levels for limiting excessive currents in a circuit in a fault condition such as an electrical short. The current limiting device comprises a magnetic core wound with two magnetically opposed, parallel connected coils of copper, a high temperature superconductor or other electrically conducting material, and a fault element connected in series with one of the coils. Under normal operating conditions, the magnetic flux density produced by the two coils cancel each other. Under a fault condition, the fault element is triggered to cause an imbalance in the magnetic flux density between the two coils which results in an increase in the impedance in the coils. While the fault element may be a separate current limiter, switch, fuse, bimetal strip or the like, it preferably is a superconductor current limiter conducting one-half of the current load compared to the same limiter wired to carry the total current of the circuit. The major voltage during a fault condition is in the coils wound on the common core in a preferred embodiment.

  6. Position measuring device

    International Nuclear Information System (INIS)

    Maeda, Kazuyuki; Takahashi, Shuichi; Maruyama, Mayumi

    1998-01-01

    The present invention provides a device capable of measuring accurate position and distance easily even at places where operator can not easily access, such as cell facilities for vitrifying radioactive wastes. Referring to a case of the vitrifying cell, an objective equipment settled in the cell is photographed by a photographing device. The image is stored in a position measuring device by way of an image input device. After several years, when the objective equipment is exchanged, a new objective equipment is photographed by a photographing device. The image is also stored in the position measuring device. The position measuring device compares the data of both of the images on the basis of pixel unit. Based on the image of the equipment before the exchange as a reference, extent of the displacement of the installation position of the equipment on the image after the exchange caused by installation error and manufacturing error is determined to decide the position of the equipment after exchange relative to the equipment before exchange. (I.S.)

  7. Temperature measuring device

    International Nuclear Information System (INIS)

    Brixy, H.

    1977-01-01

    The temperature measuring device is equipped with an electric resistor installed within a metal shroud tube so as to be insulated from it, the noise voltage of which resistor is fed to a measuring unit. The measuring junctions of one or two thermocouples are connected with the electric resistor and the legs of one or both thermocouples can be connected to the measuring unit by means of a switch. (orig.) [de

  8. Coating thickness measuring device

    International Nuclear Information System (INIS)

    Joffe, B.B.; Sawyer, B.E.; Spongr, J.J.

    1984-01-01

    A device especially adapted for measuring the thickness of coatings on small, complexly-shaped parts, such as, for example, electronic connectors, electronic contacts, or the like. The device includes a source of beta radiation and a radiation detector whereby backscatter of the radiation from the coated part can be detected and the thickness of the coating ascertained. The radiation source and detector are positioned in overlying relationship to the coated part and a microscope is provided to accurately position the device with respect to the part. Means are provided to control the rate of descent of the radiation source and radiation detector from its suspended position to its operating position and the resulting impact it makes with the coated part to thereby promote uniformity of readings from operator to operator, and also to avoid excessive impact with the part, thereby improving accuracy of measurement and eliminating damage to the parts

  9. Radiation flux measuring device

    International Nuclear Information System (INIS)

    Corte, E.; Maitra, P.

    1977-01-01

    A radiation flux measuring device is described which employs a differential pair of transistors, the output of which is maintained constant, connected to a radiation detector. Means connected to the differential pair produce a signal representing the log of the a-c component of the radiation detector, thereby providing a signal representing the true root mean square logarithmic output. 3 claims, 2 figures

  10. Radiation ray measuring device

    International Nuclear Information System (INIS)

    Maekawa, Tatsuyuki; Ida, Masaki.

    1997-01-01

    The present invention provides a chained-radiation ray monitoring system which can be applied to an actual monitoring system of a nuclear power plant or the like. Namely, this device comprises a plurality of scintillation detectors. Each of the detectors has two light take-out ports for emitting light corresponding to radiation rays irradiated from the object of the measurement to optical fibers. In addition, incident light from the optical fiber by way of one of the light take-out optical ports is transmitted to the other of the ports and sent from the other optical port to the fibers. Plurality sets of measuring systems are provided in which each of the detectors are disposed corresponding to a plurality of objects to be measured. A signal processing device is (1) connected with optical fibers of plurality sets of measuring systems in conjunction, (2) detects the optical pulses inputted from the optical fibers to identify the detector from which the optical pulses are sent and (3) measures the amount of radiation rays detected by the identified detector. As a result, the device of the present invention can form a measuring system with redundancy. (I.S.)

  11. Neutron measuring device

    International Nuclear Information System (INIS)

    Hatayama, Akiyoshi; Seki, Eiji; Kita, Yoshio; Nishitani, Takeo.

    1993-01-01

    The device of the present invention concerns measurement for neutrons in a tokamak type thermonuclear device and it can measure total amount of generated neutrons accurately throughout the operation period even if an error is caused in counted values by plasma disruption. That is, the device comprises (1) a means for detecting presence or absence of occurrence of plasma disruption and the time for the initiation of the occurrence, (2) a first data processing means for processing detection signals, (3) a means for detecting neutrons generated in plasmas and (4) a second data processing means for calculating integrated values for the number of neutrons generated from the start to the completion of electric discharge when no disruption occurs and calculating integrated values for the number of generated neutrons from the start of electric discharge to the time at the initiation of occurrence of the disruption when disruption is present. In the thus constituted device, even if an error is caused by frequent occurrence of plasma disruption, total time integrated amount of neutrons generated in the plasmas can be measured accurately. (I.S.)

  12. A power measuring device

    International Nuclear Information System (INIS)

    As, R. van.

    1985-01-01

    As a part of the klystron test facility of the Dutch NIKHEF-K accelerator, a sensitive power measuring device has been built. The high-frequency power of a klystron is stored in a water-cooled dummy load. Using a microcomputer, the increase of the water temperature and the water flow rate are transformed to a digital indication of the klystron power. (Auth.)

  13. Temperature measuring device

    Energy Technology Data Exchange (ETDEWEB)

    Lauf, R.J.; Bible, D.W.; Sohns, C.W.

    1999-10-19

    Systems and methods are described for a wireless instrumented silicon wafer that can measure temperatures at various points and transmit those temperature readings to an external receiver. The device has particular utility in the processing of semiconductor wafers, where it can be used to map thermal uniformity on hot plates, cold plates, spin bowl chucks, etc. without the inconvenience of wires or the inevitable thermal perturbations attendant with them.

  14. Temperature measurement device

    International Nuclear Information System (INIS)

    Fournier, Christian; Lions, Noel.

    1975-01-01

    The present invention relates to a temperature measuring system that can be applied in particular to monitoring the temperature of the cooling liquid metal of the outlet of the core assemblies of a fast reactor. Said device combines a long hollow metallic pole, at least partially dipped into the liquid metal and constituting a first thermocouple junction between said pole, and two metallic conductors of different nature, joined at one of their ends to constitute the second thermocouple junction. Said conductors suitably insulated are arranged inside a sheath. Said sheath made of the same metals as the pole extends inside the latter and is connected with the pole through a soldered joint. Said reliable system permits an instantaneous measurement of a quantity representing the variations in the recorded temperature and a measurement of the mean surrounding temperature that can be direcly used as a reference for calibrating the first one [fr

  15. Electronic voltage and current transformers testing device.

    Science.gov (United States)

    Pan, Feng; Chen, Ruimin; Xiao, Yong; Sun, Weiming

    2012-01-01

    A method for testing electronic instrument transformers is described, including electronic voltage and current transformers (EVTs, ECTs) with both analog and digital outputs. A testing device prototype is developed. It is based on digital signal processing of the signals that are measured at the secondary outputs of the tested transformer and the reference transformer when the same excitation signal is fed to their primaries. The test that estimates the performance of the prototype has been carried out at the National Centre for High Voltage Measurement and the prototype is approved for testing transformers with precision class up to 0.2 at the industrial frequency (50 Hz or 60 Hz). The device is suitable for on-site testing due to its high accuracy, simple structure and low-cost hardware.

  16. Reactor power measuring device

    International Nuclear Information System (INIS)

    Izumi, Mikio; Sano, Yuji; Seki, Eiji; Yoshida, Toshifumi; Ito, Toshiaki.

    1993-01-01

    The present invention provides a self-powered long detector having a sensitivity over the entire length of a reactor core as an entire control rod withdrawal range of a BWR type reactor, and a reactor power measuring device using a gamma ray thermometer which scarcely causes sensitivity degradation. That is, a hollow protection pipe is disposed passing through the reactor core from the outside of a reactor pressure vessel. The self-powered long detectors and the gamma ray thermometers are inserted and installed in the protection pipe. An average reactor power in an axial direction of the reactor relative to a certain position in the horizontal cross section of the reactor core is determined based on the power of the self-powered long detector over the entire length of the reactor core. Since the response of the self-powered detector relative to a local power change is rapid, the output is used as an input signal to a safety protection device of the reactor core. Further, a gamma ray thermometer secured in the reactor and having scarce sensitivity degradation is used instead of an incore travelling neutron monitor used for relative calibration of an existent neutron monitor secured in the reactor. (I.S.)

  17. Current scaling of plasma focus devices

    International Nuclear Information System (INIS)

    Schiuma, C.; Herold, H.; Kaeppeler, H.J.; Shakhatre, M.; Auluck, S.K.H.

    1990-03-01

    In continuation of the work by G. Decker et al. on current and neutron yield scaling of plasma focus devices an analytical solution for the circuit equation (with resistance R = 0) in the compression phase was derived. Together with the solution for the rundown phase from G. Decker et al, which was extended for finite resistance (R ≠ 0), there follows an analytical scaling theory for maximum and pinch currents. At the same time there exists the possibility to discuss the influence of finite resistance on current variation and scaling parameters. The model solutions were checked out by numerical integrations of the current equation. While at the beginning of the rundown phase the ohmic resistance cannot be neglected (the magnitude R/L plays an important role), its influence at the end of the rundown phase and in the compression phase is negligible. The theoretically determined values are compared with the results of numerous probe measurements. (orig.)

  18. Reducing leakage current in semiconductor devices

    Energy Technology Data Exchange (ETDEWEB)

    Lu, Bin; Matioli, Elison de Nazareth; Palacios, Tomas Apostol

    2018-03-06

    A semiconductor device includes a first region having a first semiconductor material and a second region having a second semiconductor material. The second region is formed over the first region. The semiconductor device also includes a current blocking structure formed in the first region between first and second terminals of the semiconductor device. The current blocking structure is configured to reduce current flow in the first region between the first and second terminals.

  19. A device for measuring electron beam characteristics

    Directory of Open Access Journals (Sweden)

    M. Andreev

    2017-01-01

    Full Text Available This paper presents a device intended for diagnostics of electron beams and the results obtained with this device. The device comprises a rotating double probe operating in conjunction with an automated probe signal collection and processing system. This provides for measuring and estimating the electron beam characteristics such as radius, current density, power density, convergence angle, and brightness.

  20. Modulated Current Drive Measurements

    International Nuclear Information System (INIS)

    Petty, C.C.; Lohr, J.; Luce, T.C.; Prater, R.; Cox, W.A.; Forest, C.B.; Jayakumar, R.J.; Makowski, M.A.

    2005-01-01

    A new measurement approach is presented which directly determines the noninductive current profile from the periodic response of the motional Stark effect (MSE) signals to the slow modulation of the external current drive source. A Fourier transform of the poloidal magnetic flux diffusion equation is used to analyze the MSE data. An example of this measurement technique is shown using modulated electron cyclotron current drive (ECCD) discharges from the DIII-D tokamak

  1. Current disruption in toroidal devices

    International Nuclear Information System (INIS)

    1979-07-01

    Attempts at raising the density or the plasma current in a tokamak above certain critical values generally result in termination of the discharge by a disruption. This sudden end of the plasma current and plasma confinement is accompanied by large induced voltages and currents in the outer structures which, in large tokamaks, can only be handled with considerable effort, and which will probably only be tolerable in reactors as rare accidents. Because of its crucial importance for the construction and operation of tokamaks, this phenomenon and its theoretical interpretation were the subject of a three-day symposium organized by the International Atomic Energy Agency and Max-Planck-Institut fuer Plasmaphysik at Garching from February 14 to 16. (orig./HT)

  2. Underwater radiation measuring device

    International Nuclear Information System (INIS)

    Seki, Noriyuki; Suzuki, Yasuo

    1998-01-01

    The present invention provides a device for measuring, under water, radiation from spent fuels (long members to be detected) of nuclear power plants and reprocessing facilities. Namely, a detecting insertion tube (insertion tube) is disposed so as to be in parallel with axial direction of the long member to be detected stored underwater. A γ-ray detector is inserted to the inside of the insertion tube. A driving mechanism is disposed for moving the γ-ray detector in axial direction inside of the insertion tube. The driving mechanism preferably has a system that it moves the γ-ray detector by winding a detection signal cable around a driving drum. The driving mechanism is formed by inserting and securing a driving tube having screws formed on the side surface and inserting it into the insertion tube. It may have a system of moving the γ-ray detector together with the driving tube while engaging the teeth of a driving transfer mechanism with the screws of the driving tube. (I.S.)

  3. A comprehensive study of charge trapping in organic field-effect devices with promising semiconductors and different contact metals by displacement current measurements

    International Nuclear Information System (INIS)

    Bisoyi, Sibani; Tiwari, Shree Prakash; Rödel, Reinhold; Zschieschang, Ute; Klauk, Hagen; Kang, Myeong Jin; Takimiya, Kazuo

    2016-01-01

    A systematic and comprehensive study on the charge-carrier injection and trapping behavior was performed using displacement current measurements in long-channel capacitors based on four promising small-molecule organic semiconductors (pentacene, DNTT, C 10 -DNTT and DPh-DNTT). In thin-film transistors, these semiconductors showed charge-carrier mobilities ranging from 1.0 to 7.8 cm 2 V −1 s −1 . The number of charges injected into and extracted from the semiconductor and the density of charges trapped in the device during each measurement were calculated from the displacement current characteristics and it was found that the density of trapped charges is very similar in all devices and of the order 10 12 cm −2 , despite the fact that the four semiconductors show significantly different charge-carrier mobilities. The choice of the contact metal (Au, Ag, Cu, Pd) was also found to have no significant effect on the trapping behavior. (paper)

  4. Eddy current analysis in fusion devices

    International Nuclear Information System (INIS)

    Turner, L.R.

    1988-06-01

    In magnetic fusion devices, particularly tokamaks and reversed field pinch (RFP) experiments, time-varying magnetic fields are in intimate contact with electrically conducting components of the device. Induced currents, fields, forces, and torques result. This note reviews the analysis of eddy current effects in the following systems: Interaction of a tokamak plasma with the eddy currents in the first wall, blanket, and shield (FWBS) systems; Eddy currents in a complex but two-dimensional vacuum vessel, as in TFTR, JET, and JT-60; Eddy currents in the FWBS system of a tokamak reactor, such as NET, FER, or ITER; and Eddy currents in a RFP shell. The cited studies are chosen to be illustrative, rather than exhaustive. 42 refs

  5. Temperature measurement device

    International Nuclear Information System (INIS)

    Oltman, B.G.; Eckerman, K.F.; Romberg, G.P.; Prepejchal, W.

    1975-01-01

    Thermoluminescent dosimeter (TLD) material is exposed to a known amount of radiation and then exposed to the environment where temperature measurements are to be taken. After a predetermined time period, the TLD material is read in a known manner to determine the amount of radiation energy remaining in the TLD material. The difference between the energy originally stored by irradiation and that remaining after exposure to the temperature ofthe environment is a measure of the average temperature of the environment during the exposure. (U.S.)

  6. A probe for Eddy current inspection devices

    International Nuclear Information System (INIS)

    1974-01-01

    The invention relates to a surface probe for Eddy current inspection devices. According to the invention, said probe comprises two magnetic core windings, with their axes in parallel relationship and at right angles to the surface of the part to be inspected. This can be applied to the nondestructive inspection of reactor components [fr

  7. Dispersion and current measurements

    International Nuclear Information System (INIS)

    Boelskifte, S.

    1986-04-01

    A model for the simulation of particle movements in water should incorporate the mutual distance dependent correlation. As long as reliable data are given accessible a model can be created of the dispersion in a given area from a statistical description of turbulence. Current measurements have been performed in an area north of the Swedish nuclear power plant Barsebaeck, and statistical time series analysis have made it possible to estimate multivariate autoregressive moving-average (ARMA) models for these data using the Box-Jenkins method. The correlation structure for the area has been investigated in detail. Transport and dispersion models for the marine environment are used in estimating doses to the population from the aquatic food chain. Some of these models are described with special emphasis on the time and length scales they cover. Furthermore, to illustrate the background of the simulation model, short introductuions are given to health physics, time series analysis, and turbulence theory. Analysis of the simulation model shows the relative importance of the different parameters. The model can be expanded to conditional simulation, where the current measurements are used directly to simulate the movement of one of the particles. Results from the model are also compared to results from a sampling of bioindicators (Fucus vesiculosus) along the Danish coast. The reliability of bioindicators in this kind of experiment is discussed. (author)

  8. Scattering measurements in Tokamak type devices

    International Nuclear Information System (INIS)

    Matoba, Tohru

    1975-03-01

    Theories, experiments and proposals for light scattering in Tokamak type devices are reviewed. Thomson scattering, measuring method of the current density distribution by scattering and resonance fluorescence are summarily described. These methods may be useful for diagnosis of the fusion plasmas. The report may help planning of the measuring apparatus for the fusion plasmas in future. (auth.)

  9. Inducer Hydrodynamic Load Measurement Devices

    Science.gov (United States)

    Skelley, Stephen E.; Zoladz, Thomas F.

    2002-01-01

    Marshall Space Flight Center (MSFC) has demonstrated two measurement devices for sensing and resolving the hydrodynamic loads on fluid machinery. The first - a derivative of the six component wind tunnel balance - senses the forces and moments on the rotating device through a weakened shaft section instrumented with a series of strain gauges. This "rotating balance" was designed to directly measure the steady and unsteady hydrodynamic loads on an inducer, thereby defining both the amplitude and frequency content associated with operating in various cavitation modes. The second device - a high frequency response pressure transducer surface mounted on a rotating component - was merely an extension of existing technology for application in water. MSFC has recently completed experimental evaluations of both the rotating balance and surface-mount transducers in a water test loop. The measurement bandwidth of the rotating balance was severely limited by the relative flexibility of the device itself, resulting in an unexpectedly low structural bending mode and invalidating the higher frequency response data. Despite these limitations, measurements confirmed that the integrated loads on the four-bladed inducer respond to both cavitation intensity and cavitation phenomena. Likewise, the surface-mount pressure transducers were subjected to a range of temperatures and flow conditions in a non-rotating environment to record bias shifts and transfer functions between the transducers and a reference device. The pressure transducer static performance was within manufacturer's specifications and dynamic response accurately followed that of the reference.

  10. Multichannel measuring instrument of small currents

    International Nuclear Information System (INIS)

    Kunetsov, A.N.; Kuznetsov, E.A.

    2009-01-01

    The device intended for display of a profile of an ion beam by measurement of currents of the co-ordinate wires is developed. Technical characteristics, basic electric and logic schemes, time diagrams, and also photos of the basic parts of the device are presented

  11. Measurement technology of RF interference current in high current system

    Science.gov (United States)

    Zhao, Zhihua; Li, Jianxuan; Zhang, Xiangming; Zhang, Lei

    2018-06-01

    Current probe is a detection method commonly used in electromagnetic compatibility. With the development of power electronics technology, the power level of power conversion devices is constantly increasing, and the power current of the electric energy conversion device in the electromagnetic launch system can reach 10kA. Current probe conventionally used in EMC (electromagnetic compatibility) detection cannot meet the test requirements on high current system due to the magnetic saturation problem. The conventional high current sensor is also not suitable for the RF (Radio Frequency) interference current measurement in high current power device due to the high noise level in the output of active amplifier. In this paper, a passive flexible current probe based on Rogowski coil and matching resistance is proposed that can withstand high current and has low noise level, to solve the measurement problems of interference current in high current power converter. And both differential mode and common mode current detection can be easily carried out with the proposed probe because of the probe's flexible structure.

  12. Fixed type incore measuring device

    International Nuclear Information System (INIS)

    Oda, Naotaka; Ito, Hitoshi; Maeda, Hiroyuki

    1998-01-01

    The present invention concerns a measuring device using gamma thermometers to be used in a BWR type reactor. An input switch is inserted to the vicinity of a detection signal input portion of a signal cable connecting GT with the detection signal input portion of a fixed type incore measuring device, and a loop resistance measuring means is disposed to the input switch on the side of the GT by way of a measurement switch. Upon measuring loop resistance, the GT measuring circuit is switched from the detection signal input portion to the loop resistance measuring means by a switching operation of the input switch and the measurement switch thereby enabling to confirm the value of the loop resistance. In addition, the lowering of the voltage in the loop resistance is compensated to confirm the accurate measurement values to be used thereby enabling to measure GT detection signals accurately. A diagnosing means for diagnosing the state of GT based on the results of the measurement for the loop resistance is disposed, and the results are reported to an operator. (N.H.)

  13. Sensitive beam current measurement for FAIR

    Energy Technology Data Exchange (ETDEWEB)

    Schwickert, Marcus; Kurian, Febin; Reeg, Hansjoerg [GSI Helmholtzzentrum fuer Schwerionenforschung GmbH, Darmstadt (Germany); Seidel, Paul; Neubert, Ralf [Friedrich-Schiller-Universitaet Jena (Germany); Geithner, Rene; Vodel, Wolfgang [Helmholtz-Institut Jena (Germany)

    2012-07-01

    Presently FAIR, the Facility for Antiproton and Ion Research, entered the final planning phase at GSI. The new accelerator facility requires precise devices for beam current measurements due to the large dynamics in beam intensities for the various synchrotrons, transport lines and storage rings. We report on the actual developments of beam diagnostic devices for the measurement of beam intensities ranging from 5 x 10{sup 11} uranium ions down to the detection of less than 10{sup 4} antiprotons. This contribution gives an overview of the planned instruments with a focus on non-intercepting beam current transformers, and summarizes the on-going development of a cryogenic current comparator.

  14. Device for measuring well twistings

    Energy Technology Data Exchange (ETDEWEB)

    Kostin, Yu S; Golubin, S V; Keller, V F; Merzheyevskiy, A B; Zdorov, V P

    1982-01-01

    The device for measuring the well twistings with the use of fluids (poured into a vessel and which leave an imprint on the walls), containing a housing and adapter, is distinguished by the fact that in order to improve the accuracy of measurement by obtaining a clear imprint, it is equipped with cylinder that is spring-loaded in relation to the adapter, forming a vessel for fluid with the adapter. The adapter is made of two parts, one of which is made of neutral metal in relation to the fluid, and the other, from active in relation to the same fluid.

  15. Automatic adjustment of bias current for direct current superconducting quantum interference device

    International Nuclear Information System (INIS)

    Makie-Fukuda, K.; Hotta, M.; Okajima, K.; Kado, H.

    1993-01-01

    A new method of adjusting the bias current of dc superconducting quantum interference device (SQUID) is described. It is shown that the signal-to-noise ratio of a SQUID magnetometer connected in a flux-locked loop configuration is proportional to the second harmonic of the output signal from the SQUID. A circuit configuration that can automatically optimize a SQUID's bias current by measuring this second harmonic and adjusting the bias current accordingly is proposed

  16. Radiation dose rate measuring device

    International Nuclear Information System (INIS)

    Sorber, R.

    1987-01-01

    A portable device is described for in-field usage for measuring the dose rate of an ambient beta radiation field, comprising: a housing, substantially impervious to beta radiation, defining an ionization chamber and having an opening into the ionization chamber; beta radiation pervious electrically-conductive window means covering the opening and entrapping, within the ionization chamber, a quantity of gaseous molecules adapted to ionize upon impact with beta radiation particles; electrode means disposed within the ionization chamber and having a generally shallow concave surface terminating in a generally annular rim disposed at a substantially close spacing to the window means. It is configured to substantially conform to the window means to define a known beta radiation sensitive volume generally between the window means and the concave surface of the electrode means. The concave surface is effective to substantially fully expose the beta radiation sensitive volume to the radiation field over substantially the full ambient area faced by the window means

  17. New devices for radon measurements

    International Nuclear Information System (INIS)

    Sevostyanov, V.N.

    2004-01-01

    This work includes the description of two new devices for radon surveys developed by the authors and produced in Kazakhstan. The first appliance is 'Ramon-Radon-01' used to measure 222 Rn radon in various mediums such as air, water, soil, and radon exhalation. The major advantage of the appliance lies in the absence of radioactive pollution in it after measurements. The appliances widely used in the CIS such as 'RAA-01', 'Alpharad' (produced by 'MTM Zaschita', Russia) and 'Alphaguard' (Germany) take samples directly to the measuring camera. For instance, the activity concentration of samples after they are taken by 'RAA-01' and 'Alpharad' is measured by means of electrostatic precipitation of RaA ( 218 Po) atoms to the square of semiconductor detector with subsequent registration of RaA alpha decay. The obvious disadvantage is that the subsequent measurement of relatively small 222 Rn activity concentration values after great values of 222 Rn activity concentration have been obtained requires a considerable exposure of the appliance sometimes exceeding 10 hours. Therefore, appliances register a relatively low value of the top measurement range of 20 KBq/m 3 . 'Alphaguard' has similar limitation resulting from precipitation of radon daughter decay products on the walls of ionizing chamber where radon activity concentration is measured. The radioactive lag of 'RAA-01', 'Alpharad' and 'Alphaguard' makes them of little use as well for automatic monitoring in the conditions of abruptly time negative derivatives on change of radon activity concentration. The second advantage is that 'Ramon-Radon-01', as opposed to above described appliances, registers almost zero radioactive lag, thanks to its constructive peculiarities which enable an abrupt increase of top range of measured value up to 5x10 5 Bq/m 3 , only limited by velocity of electron units of the appliance. The third advantage is that measurement discontinuity is determined only by time of full measurement cycle

  18. Current voltage perspective of an organic electronic device

    Science.gov (United States)

    Mukherjee, Ayash K.; Kumari, Nikita

    2018-05-01

    Nonlinearity in current (I) - voltage (V) measurement is a well-known attribute of two-terminal organic device, irrespective of the geometrical or structural arrangement of the device. Most of the existing theories that are developed for interpretation of I-V data, either focus current-voltage relationship of charge injection mechanism across the electrode-organic material interface or charge transport mechanism through the organic active material. On the contrary, both the mechanisms work in tandem charge conduction through the device. The transport mechanism is further complicated by incoherent scattering from scattering centres/charge traps that are located at the electrode-organic material interface and in the bulk of organic material. In the present communication, a collective expression has been formulated that comprises of all the transport mechanisms that are occurring at various locations of a planar organic device. The model has been fitted to experimental I-V data of Au/P3HT/Au device with excellent degree of agreement. Certain physical parameters such as the effective area of cross-section and resistance due to charge traps have been extracted from the fit.

  19. Reactor power region measuring device

    International Nuclear Information System (INIS)

    Kashiwa, Takao.

    1996-01-01

    The device of the present invention can rapidly detect abnormality of a local power region monitor (LPRM) even at a low power region caused such as upon start-up of a BWR type reactor. Namely, the present invention comprises (1) an LPRM detector for measuring neutron fluxes in the reactor, (2) a gamma thermo detector for calibrating the sensitivity of the LPRM detector, (3) a comparison circuit for comparing the detected values of the detectors (1) and (2), and (4) an alarm circuit for outputting an alarm when the comparative difference of the output of the circuit (3) exceeds a predetermined value. Signals of an alarm for a lower limit of the LPRM detector have been issued continuously upon start-up and shut down of the reactor since neutron fluxes in the reactor are reduced. However, the gamma thermo detector is always secured in the inside of the reactor different from a travelling-type incore probe monitor (TIP) disposed so far for the same purpose. Accordingly, the alarm generated upon usual start-up can be eliminated by comparing the detected values of the detector (2) and abnormality of the detector (1) can be rapidly detected by judging the abnormality of the comparative difference. (I.S.)

  20. Beam diagnostics using an emittance measurement device

    International Nuclear Information System (INIS)

    Sarstedt, M.; Becker, R.; Klein, H.; Maaser, A.; Mueller, J.; Thomae, R.; Weber, M.

    1995-01-01

    For beam diagnostics aside from Faraday cups for current measurements and analysing magnets for the determination of beam composition and energy the most important tool is an emittance measurement device. With such a system the distribution of the beam particles in phase-space can be determined. This yields information not only on the position of the particles but also on their angle with respect to the beam axis. There are different kinds of emittance measurement devices using either circular holes or slits for separation of part of the beam. The second method (slit-slit measurement), though important for the determination of the rms-emittance, has the disadvantage of integrating over the y- and y'-coordinate (measurement in xx'-plane assumed). This leads to different emittance diagrams than point-point measurements, since in xx'-plane for each two corresponding points of rr'-plane there exists a connecting line. With regard to beam aberrations this makes xx'-emittances harder to interpret. In this paper the two kinds of emittance diagrams are discussed. Additionally the influence of the slit height on the xx'-emittance is considered. The analytical results are compared to experimental measurements in rr'-, rx'- and xx'-phase-space. (orig.)

  1. Medical devices for the anesthetist: current perspectives

    Directory of Open Access Journals (Sweden)

    Ingrande J

    2014-03-01

    Full Text Available Jerry Ingrande, Hendrikus JM LemmensDepartment of Anesthesiology, Perioperative and Pain Medicine, Stanford University School of Medicine, Stanford, CA, USAAbstract: Anesthesiologists are unique among most physicians in that they routinely use technology and medical devices to carry out their daily activities. Recently, there have been significant advances in medical technology. These advances have increased the number and utility of medical devices available to the anesthesiologist. There is little doubt that these new tools have improved the practice of anesthesia. Monitoring has become more comprehensive and less invasive, airway management has become easier, and placement of central venous catheters and regional nerve blockade has become faster and safer. This review focuses on key medical devices such as cardiovascular monitors, airway equipment, neuromonitoring tools, ultrasound, and target controlled drug delivery software and hardware. This review demonstrates how advances in these areas have improved the safety and efficacy of anesthesia and facilitate its administration. When applicable, indications and contraindications to the use of these novel devices will be explored as well as the controversies surrounding their use.Keywords: catheters, echocardiography, ultrasound, fiberoptic bronchoscope, laryngeal mask airway, closed-loop anesthesia

  2. An application of residual current protective device at electrical installation

    International Nuclear Information System (INIS)

    Firman Silitonga

    2008-01-01

    In an electrical installation, a protection for overload and short circuit are always to be installed. In addition to the installation, it is necessary to be installed a protection device for residual current because both the short circuit and the overload device protection will not work for the residual current. The quantity of the residual current must be defined first at any electrical installation to define an appropriate residual current protection so that not every residual current will break the circuit down. This paper will explain a method how to install a residual protection device for 3500 VA or more at TN and TT of earthing system. (author)

  3. Eddy current testing device using unbalance bridge

    International Nuclear Information System (INIS)

    Hoshikawa, H.; Koido, J.; Ishibashi, Y.

    1976-01-01

    An easily readjustable unbalance bridge has been invented and in utilizing the same, an eddy current testing equipment excellent in suppression of the lift-off effect and high in the detection sensitivity has been developed

  4. Reactor water level measuring device

    International Nuclear Information System (INIS)

    Kuroki, Reiji; Asano, Tamotsu.

    1996-01-01

    A condensation vessel is connected to the upper portion of a reactor pressure vessel by way of a pipeline. The lower portion of the condensation vessel is connected to a low pressure side of a differential pressure transmission device by way of a reference leg pipeline. The high pressure side of the differential pressure transmission device is connected to the lower portion of the pressure vessel by way of a pipeline. The condensation vessel is equipped with a temperature sensor. When a temperature of a gas phase portion in the condensation vessel is lowered below a predetermined level, and incondensible gases in the condensation vessel starts to be dissolved in water, signals are sent from the temperature sensor to a control device and a control valve is opened. With such a constitution, CRD driving water flows into the condensation vessel, and water in which gases at the upper portion of the condensation vessel is dissolved flows into the pressure vessel by way of a pipeline. Then, gases dissolved in a reference water column in the reference leg pipeline are eliminated and the value of a reference water pressure does not change even upon abrupt lowering of pressure. (I.N.)

  5. Method and device for measuring fluid flow

    International Nuclear Information System (INIS)

    Atherton, R.; Marinkovich, P.S.; Spadaro, P.R.; Stout, J.W.

    1976-01-01

    The invention is a fluid flow measuring device for determining the coolant flow at the entrance to a specific nuclear reactor fuel region. The device comprises a plurality of venturis having the upstream inlet and throat pressure of each respectively manifolded together to provide one static pressure signal for each region monitored. The device provides accurate flow measurement with low pressure losses and uniform entrance and discharge flow distribution. 1 claim, 7 figures

  6. New Devices for Electret Measurement

    Directory of Open Access Journals (Sweden)

    ROMANIUC Ilie

    2013-05-01

    Full Text Available Green energy research have increased in recent decades and interest in unconventional energy production systems is becoming more pronounced. Such unconventional energy generation system is alsobased on electret generator. The electret, is a dielectric material that has a quasi-permanent electric charge. This paper, provides a brief history of the Carnauba wax electret, will be described the process of electret making and will be presented two devices formeasuring the carnauba wax electrets, one of which uses magnetostrictive phenomenon and the second design is achieved with an oscilomotor.

  7. A measurement device for electromagnetic flow tomography

    Science.gov (United States)

    Vauhkonen, M.; Hänninen, A.; Lehtikangas, O.

    2018-01-01

    Electromagnetic flow meters have succesfully been used in many industries to measure the mean flow velocity of conductive liquids. This technology works reliably in single phase flows with axisymmetric flow profiles but can be inaccurate with asymmetric flows, which are encountered, for example, in multiphase flows, pipe elbows and T-junctions. Some computational techniques and measurement devices with multiple excitation coils and measurement electrodes have recently been proposed to be used in cases of asymmetric flows. In earlier studies, we proposed a computational approach for electromagnetic flow tomography (EMFT) for estimating velocity fields utilizing several excitation coils and a set of measurement electrodes attached to the surface of the pipe. This approach has been shown to work well with simulated data but has not been tested extensively with real measurements. In this paper, an EMFT system with four excitation coils and 16 measurement electrodes is introduced. The system is capable of using both square wave and sinusoidal coil current excitations and all the coils can be excited individually, also enabling parallel excitations with multiple frequencies. The studies undertaken in the paper demonstrate that the proposed EMFT system, together with the earlier introduced velocity field reconstruction approach, is capable of producing reliable velocify field estimates in a laboratory environment with both axisymmetric and asymmetric single phase flows.

  8. Measurement-device-independent quantum digital signatures

    Science.gov (United States)

    Puthoor, Ittoop Vergheese; Amiri, Ryan; Wallden, Petros; Curty, Marcos; Andersson, Erika

    2016-08-01

    Digital signatures play an important role in software distribution, modern communication, and financial transactions, where it is important to detect forgery and tampering. Signatures are a cryptographic technique for validating the authenticity and integrity of messages, software, or digital documents. The security of currently used classical schemes relies on computational assumptions. Quantum digital signatures (QDS), on the other hand, provide information-theoretic security based on the laws of quantum physics. Recent work on QDS Amiri et al., Phys. Rev. A 93, 032325 (2016);, 10.1103/PhysRevA.93.032325 Yin, Fu, and Zeng-Bing, Phys. Rev. A 93, 032316 (2016), 10.1103/PhysRevA.93.032316 shows that such schemes do not require trusted quantum channels and are unconditionally secure against general coherent attacks. However, in practical QDS, just as in quantum key distribution (QKD), the detectors can be subjected to side-channel attacks, which can make the actual implementations insecure. Motivated by the idea of measurement-device-independent quantum key distribution (MDI-QKD), we present a measurement-device-independent QDS (MDI-QDS) scheme, which is secure against all detector side-channel attacks. Based on the rapid development of practical MDI-QKD, our MDI-QDS protocol could also be experimentally implemented, since it requires a similar experimental setup.

  9. Improvements in backscatter measurement devices

    International Nuclear Information System (INIS)

    Saunders, J.; Hay, W.D.

    1978-01-01

    Improvements in measuring the thickness of a coating on a substrate by the technique of backscattered particles are described. These improvements enable the measurements to be carried out continuously as an integral part of the coating production line and also permit measurements where the coated elements are separated from one another by a predetermined distance. The former is achieved by situating the backscatter probe and detector on the rim of the measurement wheel and rotating this wheel at a speed such that the coated element and probe are stationary relative to one another. The latter improvement is achieved by an indexing apparatus which automatically positions the probe beside a coated element. (U.K.)

  10. ALS insertion device block measurement and inspection

    International Nuclear Information System (INIS)

    Marks, S.; Carrieri, J.; Cook, C.; Hassenzahl, W.V.; Hoyer, E.; Plate, D.

    1991-05-01

    The performance specifications for ALS insertion devices require detailed knowledge and strict control of the Nd-Fe-B permanent magnet blocks incorporated in these devices. This paper describes the measurement and inspection apparatus and the procedures designed to qualify and characterize these blocks. A detailed description of a new, automated Helmholtz coil facility for measurement of the three components of magnetic moment is included. Physical block inspection and magnetic moment measurement procedures are described. Together they provide a basis for qualifying blocks and for specifying placement of blocks within an insertion devices' magnetic structures. 1 ref., 4 figs

  11. Dark Current Measurements in GIF++

    CERN Document Server

    Al-Qahtani, Abdulaziz

    2017-01-01

    This project revolved around creating a code that treats experiment files in order to fetch dark current measurements done for the Cathode Strip Chambers (CSC) exposed to long-term irradiation at Gamma Irradiation Facility (GIF++) and plot the results as functions of the accumulated charge. This paper will discuss the Compact Muon Solenoid (CMS) experiment, the CSCs that we have looked at (specifically the muon detectors located at the CMS endcaps), longevity test at GIF++, and finally discuss the code and results obtained.

  12. Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer.

    Science.gov (United States)

    Chen, Yen-An; Hirnschall, Nino; Findl, Oliver

    2011-03-01

    To compare 2 new optical biometry devices with the present gold standard biometer. Vienna Institute for Research in Ocular Surgery, Department of Ophthalmology, Hanusch Hospital, Vienna, Austria. Evaluation of diagnostic test or technology. In patients scheduled for cataract surgery, measurements performed with the current gold standard optical biometer (IOLMaster) were compared with those of 2 new optical biometers, the Lenstar LS 900 (optical low-coherence reflectometry [OLCR] device; substudy 1) and the IOLMaster 500 (partial coherence interferometry [PCI] device; substudy 2). The duration of patient data entry and of the actual measurement process and the time from intraocular lens power calculation to printout were calculated. The mean difference in axial length measurements was 0.01 mm ± 0.05 (SD) between the gold standard device and the new OLCR device and 0.01 ± 0.02 mm between the gold standard device and the new PCI device (P=.12 and P gold standard device (mean difference 209 ± 127 seconds), and measurements with the gold standard device took significantly longer than with the new PCI device (mean difference 82 ± 46 seconds) (both P gold standard device. Measurements with the new OLCR device took twice as long as those with the gold standard device. Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

  13. A device for electron gun emittance measurement

    International Nuclear Information System (INIS)

    Aune, B.; Corveller, P.; Jablonka, M.; Joly, J.M.

    1985-05-01

    In order to improve the final emittance of the beam delivered by the ALS electron linac a new gun is going to be installed. To measure its emittance and evaluate the contribution of different factors to emittance growth we have developed an emittance measurement device. We describe the experimental and mathematical procedure we have followed, and give some results of measurements

  14. Measurement device-independent quantum dialogue

    Science.gov (United States)

    Maitra, Arpita

    2017-12-01

    Very recently, the experimental demonstration of quantum secure direct communication (QSDC) with state-of-the-art atomic quantum memory has been reported (Zhang et al. in Phys Rev Lett 118:220501, 2017). Quantum dialogue (QD) falls under QSDC where the secrete messages are communicated simultaneously between two legitimate parties. The successful experimental demonstration of QSDC opens up the possibilities for practical implementation of QD protocols. Thus, it is necessary to analyze the practical security issues of QD protocols for future implementation. Since the very first proposal for QD by Nguyen (Phys Lett A 328:6-10, 2004), a large number of variants and extensions have been presented till date. However, all of those leak half of the secret bits to the adversary through classical communications of the measurement results. In this direction, motivated by the idea of Lo et al. (Phys Rev Lett 108:130503, 2012), we propose a measurement device-independent quantum dialogue scheme which is resistant to such information leakage as well as side-channel attacks. In the proposed protocol, Alice and Bob, two legitimate parties, are allowed to prepare the states only. The states are measured by an untrusted third party who may himself behave as an adversary. We show that our protocol is secure under this adversarial model. The current protocol does not require any quantum memory, and thus, it is inherently robust against memory attacks. Such robustness might not be guaranteed in the QSDC protocol with quantum memory (Zhang et al. 2017).

  15. Oscillation measuring device for body of rotation

    International Nuclear Information System (INIS)

    Komita, Hideo.

    1994-01-01

    The present invention concerns an internal pump of a BWR type reactor and provides a device for detecting oscillations of a rotational shaft. Namely, recesses are formed along an identical circumference on the outer circumferential surface of the rotating portion each at a predetermined distance. The recesses rotate along with the rotation. An eddy current type displacement gage measures the distance to the outer circumferential surface of the rotating portion. The recesses are detected by the displacement gage as pulse signals. When the rotating portion oscillates, it is detected by the displacement gage as waveform signals. Accordingly, the output signals of the eddy current type displacement gage are formed by pulse signals superposed on the waveform signals. A rising detection circuit detects the rising position of the pulse signals as the components of the number of rotation of the rotating portion, and fall detection circuit detects the falling position. A comparator circuit is disposed in parallel with both of rising/falling detection circuits. A predetermined threshold value is set in the comparator circuit to output a signal when the inputted signal exceeds the value. (I.S.)

  16. Dark current of organic heterostructure devices with insulating spacer layers

    Science.gov (United States)

    Yin, Sun; Nie, Wanyi; Mohite, Aditya D.; Saxena, Avadh; Smith, Darryl L.; Ruden, P. Paul

    2015-03-01

    The dark current density at fixed voltage bias in donor/acceptor organic planar heterostructure devices can either increase or decrease when an insulating spacer layer is added between the donor and acceptor layers. The dominant current flow process in these systems involves the formation and subsequent recombination of an interfacial exciplex state. If the exciplex formation rate limits current flow, the insulating interface layer can increase dark current whereas, if the exciplex recombination rate limits current flow, the insulating interface layer decreases dark current. We present a device model to describe this behavior and illustrate it experimentally for various donor/acceptor systems, e.g. P3HT/LiF/C60.

  17. Device for radioactivity measurement of liquid samples

    International Nuclear Information System (INIS)

    Lamaziere, J.

    1983-01-01

    The device for low activity gamma measurements comprises an automatic changer for sample transfer from a conveyor to a measuring chamber. The conveyor includes a horizontal table were are regularly distributed sample holders. A lift allows a vertical motion of a plate for the exposition in front of a detector [fr

  18. Device measures static friction of magnetic tape

    Science.gov (United States)

    Cole, P. T.

    1967-01-01

    Device measures the coefficient of static friction of magnetic tape over a range of temperatures and relative humidities. It uses a strain gage to measure the force of friction between a reference surface and the tape drawn at a constant velocity of approximately 0.0001 inch per second relative to the reference surface.

  19. Microstructure cantilever beam for current measurement

    Directory of Open Access Journals (Sweden)

    M.T.E. Khan

    2010-01-01

    Full Text Available Most microelectromechanical systems (MEMS sensors are based on the microcantilever technology, which uses a broad range of design materials and structures. The benefit ofMEMStechnology is in developing devices with a lower cost, lower power consumption, higher performance and greater integration. A free-end cantilever beam with a magnetic material mass has been designed using MEMS software tools. The magnetic material was used to improve the sensitivity of the cantilever beam to an externally-applied magnetic field. The cantilever was designed to form a capacitance transducer, which consisted of variable capacitance where electrical and mechanical energies were exchanged. The aim of this paper was to analyse the system design of the microcantilever when subjected to a magnetic field produced by a current-carrying conductor. When the signal, a sinusoidal current with a constant frequency, was applied, the cantilever beam exhibited a vibration motion along the vertical axis when placed closer to the line current. This motion created corresponding capacitance changes and generated a voltage output proportional to the capacitive change in the signal-processing circuitry attached to the microcantilever. The equivalent massspring system theory was used to describe and analyse the effect of the natural frequency of the system vibrations and motion due to the applied magnetic field, in a single-degree of freedom. The main application of this microcantilever is in current measurements to develop a non-contact current sensor mote.

  20. Current-ripple effect on superconductive dc critical current measurements

    International Nuclear Information System (INIS)

    Goodrich, L.F.; Bray, S.L.; Clark, A.F.

    1988-01-01

    The effect of sample-current power-supply ripple on dc critical current measurement in multifilamentary NbTi superconductors was evaluated. In general the ripple in a current supply became more significant above 500 A because effective filtering was hard to achieve. Ripple also caused noise at the input of the voltmeter used for the measurements. The quantitative effect of current ripple was studied using a battery current supply modified to allow the creation of ripple current with variable frequency and amplitude. Problems common to all large-conductor critical current measurements are discussed

  1. A comparison of 'radon' measurement devices

    International Nuclear Information System (INIS)

    Rolle, R.

    2004-01-01

    For the indoor concentration measurement of Rn decay chain members, instruments ranging from long-term integrating passive devices to far more sensitive active devices with good time resolution are in use. The former are used mainly to screen for potentially high exposure locations, while the latter are useful for assessing the exposure dynamics of encumbered premises, with potential clarification (modelling) for optimal remediation, before, during and after such process. Concentration measurement of any one decay chain member always involves its complete decay chain. The 222 Rn gas concentration can be measured with gas measuring devices or with spectrometric measuring instruments that concentrate the decay products. The latter generally offer far more accurate measurement per time of the 222 Rn concentration, than mere Rn gas measuring devices, and also give the decay product concentrations and thus activity ratios which are related to air exchange. The concentrations of the unattached Rn decay products may be measured simultaneously with inline screen measurement. In premises steady state ventilation conditions are rare. Rapidly changing concentrations can be measured more accurately with the more sensitive concentrating instruments - the inherent instrumental time lag of detected signals from chain decay, relative to time of sampling, should however be corrected where rates of concentration change approximate chain decay constants. Counting of beta-signals in addition to alpha-spectrometry, and quasi-continuous sampling while continuously measuring, enhances the sensitivity, and particularly measurement time lag correction, where concentration ratios are sought for elucidating air exchange. Appropriate software ought to be made available to evaluate the spectrometric data and to link it to suitable compartment ventilation models (automatic adjustment for a dynamic change in alpha-calibration could be incorporated) - this would constitute the modern version of

  2. Experimental Measurement-Device-Independent Entanglement Detection

    Science.gov (United States)

    Nawareg, Mohamed; Muhammad, Sadiq; Amselem, Elias; Bourennane, Mohamed

    2015-02-01

    Entanglement is one of the most puzzling features of quantum theory and of great importance for the new field of quantum information. The determination whether a given state is entangled or not is one of the most challenging open problems of the field. Here we report on the experimental demonstration of measurement-device-independent (MDI) entanglement detection using witness method for general two qubits photon polarization systems. In the MDI settings, there is no requirement to assume perfect implementations or neither to trust the measurement devices. This experimental demonstration can be generalized for the investigation of properties of quantum systems and for the realization of cryptography and communication protocols.

  3. Device for ultrasonic and eddy current testing of bolts

    International Nuclear Information System (INIS)

    Hromek, J.; Kaspar, P.

    1989-01-01

    The device provides pivoting fitting of the bolt of a WWER reactor steam generator while ultrasonic and eddy current probes are brought near. The bolt under study is clamped between a drive funnel and a securing cone. The eddy current probes are adjusted using guide arms to the point requested and are fitted over the bolt such as for their thread segments to engage the bolt thread. The ultrasonic transducers are then adjusted to the required point. The device can be used for testing bolts of a thread size from M54x5 and a maximum length of 600 mm. (J.B.). 1 fig

  4. Development of induction current acquisition device based on ARM

    Science.gov (United States)

    Ji, Yanju; Liu, Xiyang; Huang, Wanyu; Yao, Jiang; Yuan, Guiyang; Hui, Luan; Guan, Shanshan

    2018-03-01

    We design an induction current acquisition device based on ARM in order to realize high resolution and high sampling rate of acquisition for the induction current in wire-loop. Considering its characteristics of fast attenuation and small signal amplitude, we use the method of multi-path fusion for noise suppression. In the paper, the design is carried out from three aspects of analog circuit and device selection, independent power supply structure and the electromagnetic interference suppression of high frequency. DMA and ping-pong buffer, as a new data transmission technology, solves real time storage problem of massive data. The performance parameters of ARM acquisition device are tested. The comparison test of ARM acquisition device and cRIO acquisition device is performed at different time constants. The results show that it has 120dB dynamic range, 47kHz bandwidth, 96kHz sampling rate, 5μV the smallest resolution, and its average error value is not more than 4%, which proves the high accuracy and stability of the device.

  5. Current-driven dynamics in molecular-scale devices

    International Nuclear Information System (INIS)

    Seideman, Tamar

    2003-01-01

    We review recent theoretical work on current-triggered processes in molecular-scale devices - a field at the interface between solid state physics and chemical dynamics with potential applications in diverse areas, including artificial molecular machines, unimolecular transport, surface nanochemistry and nanolithography. The qualitative physics underlying current-triggered dynamics is first discussed and placed in context with several well-studied phenomena with which it shares aspects. A theory for modelling these dynamics is next formulated within a time-dependent scattering approach. Our end result provides useful insight into the system properties that determine the reaction outcome as well as a computationally convenient framework for numerical realization. The theory is applied to study single-molecule surface reactions induced by a scanning tunnelling microscope and current-triggered dynamics in single-molecule transistors. We close with a discussion of several potential applications of current-induced dynamics in molecular devices and several opportunities for future research. (topical review)

  6. Using Teacher-Made Measurement Devices.

    Science.gov (United States)

    Mehrens, William A.; Lehmann, Irvin J.

    1987-01-01

    Classroom measurement devices, when tailored to fit a teacher's particular instructional objectives, are essential for optimal teaching and learning. Teachers use test data to assess students' progress but often fail to analyze tests for validity. This article shows how locally and correctly developed assessment tools may serve a variety of…

  7. Two devices for atmospheric electric field measurement

    International Nuclear Information System (INIS)

    Colombet, Andre; Hubert, Pierre.

    1977-02-01

    Two instruments installed at St Privat d'Allier for electric field measurement in connection with the rocket triggered lighting experiment program are described. The first one is a radioactive probe electrometer used as a warning device. The second is a field mill used for tape recording of electric field variation during the triggering events. Typical examples of such records are given [fr

  8. Measuring ionizing radiation with a mobile device

    Science.gov (United States)

    Michelsburg, Matthias; Fehrenbach, Thomas; Puente León, Fernando

    2012-02-01

    In cases of nuclear disasters it is desirable to know one's personal exposure to radioactivity and the related health risk. Usually, Geiger-Mueller tubes are used to assess the situation. Equipping everyone with such a device in a short period of time is very expensive. We propose a method to detect ionizing radiation using the integrated camera of a mobile consumer device, e.g., a cell phone. In emergency cases, millions of existing mobile devices could then be used to monitor the exposure of its owners. In combination with internet access and GPS, measured data can be collected by a central server to get an overview of the situation. During a measurement, the CMOS sensor of a mobile device is shielded from surrounding light by an attachment in front of the lens or an internal shutter. The high-energy radiation produces free electrons on the sensor chip resulting in an image signal. By image analysis by means of the mobile device, signal components due to incident ionizing radiation are separated from the sensor noise. With radioactive sources present significant increases in detected pixels can be seen. Furthermore, the cell phone application can make a preliminary estimate on the collected dose of an individual and the associated health risks.

  9. Use of high current density superconducting coils in fusion devices

    International Nuclear Information System (INIS)

    Green, M.A.

    1979-11-01

    Superconducting magnets will play an important role in fusion research in years to come. The magnets which are currently proposed for fusion research use the concept of cryostability to insure stable operation of the superconducting coils. This paper proposes the use of adiabatically stable high current density superconducting coils in some types of fusion devices. The advantages of this approach are much lower system cold mass, enhanced cryogenic safety, increased access to the plasma and lower cost

  10. Temperature measurement with industrial color camera devices

    Science.gov (United States)

    Schmidradler, Dieter J.; Berndorfer, Thomas; van Dyck, Walter; Pretschuh, Juergen

    1999-05-01

    This paper discusses color camera based temperature measurement. Usually, visual imaging and infrared image sensing are treated as two separate disciplines. We will show, that a well selected color camera device might be a cheaper, more robust and more sophisticated solution for optical temperature measurement in several cases. Herein, only implementation fragments and important restrictions for the sensing element will be discussed. Our aim is to draw the readers attention to the use of visual image sensors for measuring thermal radiation and temperature and to give reasons for the need of improved technologies for infrared camera devices. With AVL-List, our partner of industry, we successfully used the proposed sensor to perform temperature measurement for flames inside the combustion chamber of diesel engines which finally led to the presented insights.

  11. Pressure measurements in magnetic-fusion devices

    International Nuclear Information System (INIS)

    Dylla, H.F.

    1981-11-01

    Accurate pressure measurements are important in magnetic fusion devices for: (1) plasma diagnostic measurements of particle balance and ion temperature; (2) discharge cleaning optimization; (3) vacuum system performance; and (4) tritium accountability. This paper reviews the application, required accuracy, and suitable instrumentation for these measurements. Demonstrated uses of ionization-type and capacitance-diaphragm gauges for various pressure and gas-flow measurements in tokamaks are presented, with specific reference to the effects of magnetic fields on gauge performance and the problems associated with gauge calibration

  12. Pressure measurements in magnetic-fusion devices

    Energy Technology Data Exchange (ETDEWEB)

    Dylla, H.F.

    1981-11-01

    Accurate pressure measurements are important in magnetic fusion devices for: (1) plasma diagnostic measurements of particle balance and ion temperature; (2) discharge cleaning optimization; (3) vacuum system performance; and (4) tritium accountability. This paper reviews the application, required accuracy, and suitable instrumentation for these measurements. Demonstrated uses of ionization-type and capacitance-diaphragm gauges for various pressure and gas-flow measurements in tokamaks are presented, with specific reference to the effects of magnetic fields on gauge performance and the problems associated with gauge calibration.

  13. Device for measuring fission product density

    International Nuclear Information System (INIS)

    Kaneda, Mitsunori.

    1980-01-01

    Purpose: To determine the fission product density of xenon or the like and enable measurement of real time of fission product density in a reactor by calculating the disintegration and annihilation of the fission product on the basis of neutron detected output. Constitution: The neutron flux in a reactor is detected by a detector, and applied to first and second density calculators. Second fission product density signal of xenon or the like outputted from first device is again inputted to the device to form an annihilation signal due to disintegration to determine the present density of the second fission product of xenon or the like corresponding to the decrease of the neutron due to the poison of xeron or the like. Similarly, second device determines the first fission product density of iodine or the like. (Sekiya, K.)

  14. Resolving Overlimiting Current Mechanisms in Microchannel-Nanochannel Interface Devices

    Science.gov (United States)

    Yossifon, Gilad; Leibowitz, Neta; Liel, Uri; Schiffbauer, Jarrod; Park, Sinwook

    2015-11-01

    We present results demonstrating the space charge-mediated transition between classical, diffusion-limited current and surface-conduction dominant over-limiting currents in a shallow micro-nanochannel device. The extended space charge layer develops at the depleted micro-nanochannel entrance at high current and is correlated with a distinctive maximum in the dc resistance. Experimental results for a shallow surface-conduction dominated system are compared with theoretical models, allowing estimates of the effective surface charge at high voltage to be obtained. Further, we extend the study to microchannels of moderate to large depths where the role of various electro-convection mechanisms becomes dominant. In particular, electro-osmotic of the second kind and electro-osmotic instability (EOI) which competes each other at geometrically heterogeneous (e.g. undulated nanoslot interface, array of nanoslots) nanoslot devices. Also, these effects are also shown to be strongly modulated by the non-ideal permselectivity of the nanochannel.

  15. Device for measuring atmospheric radon activity

    International Nuclear Information System (INIS)

    Deml, F.; Jansky, Z.; Smejkal, Z.

    1989-01-01

    The device consists of a lightproof case pivoted on a stand. Minimally two chambers are provided in the case for holding glass scintillation chambers, each with an opening in its bottom. The centers of the openings lie on a common circle with its center in the center of rotation of the case. An opening for the passage of light is provided in the stand. The opening also is on the common circle of the centres of the openings. A photomultiplier is placed below the opening. Measurement always proceeds with one scintillation chamber only. Thus, replacement of scintillation chambers can take place simultaneously with measurement and rapid and continuous measurement can be secured. Case rotation can be controlled automatically. The device can be used not only in monitoring air pollution but also in monitoring solid materials, aquifers, mine corridors, natural materials, etc. (J.B.). 1 fig

  16. Current adaptation measures and policies

    Science.gov (United States)

    Geoff Roberts; John A. Parrotta; Anita. Wreford

    2009-01-01

    As stated in earlier chapters, the possible impacts of climate change on forests and the forest sector are considerable, and many impacts have already been observed. As forest conditions change, there is an inherent need to change management and policy measures to minimise negative impacts and to exploit the benefits derived from climate change. This chapter highlights...

  17. Measuring device for control rod driving time

    International Nuclear Information System (INIS)

    Tanaka, Kazuhiko; Hanabusa, Masatoshi.

    1993-01-01

    The present invention concerns a measuring device for control driving time having a function capable of measuring a selected control rod driving time and measuring an entire control rod driving time simultaneously. A calculation means and a store means for the selected rod control rod driving time, and a calculation means and a store means for the entire control rod driving time are disposed individually. Each of them measures the driving time and stores the data independent of each other based on a selected control rod insert ion signal and an entire control rod insertion signal. Even if insertion of selected and entire control rods overlaps, each of the control rod driving times can be measured reliably to provide an advantageous effect capable of more accurately conducting safety evaluation for the nuclear reactor based on the result of the measurement. (N.H.)

  18. Improvised explosive devices: pathophysiology, injury profiles and current medical management.

    Science.gov (United States)

    Ramasamy, A; Hill, A M; Clasper, J C

    2009-12-01

    The improvised explosive device (IED), in all its forms, has become the most significant threat to troops operating in Afghanistan and Iraq. These devices range from rudimentary home made explosives to sophisticated weapon systems containing high-grade explosives. Within this broad definition they may be classified as roadside explosives and blast mines, explosive formed pojectile (EFP) devices and suicide bombings. Each of these groups causeinjury through a number of different mechanisms and can result in vastly different injury profiles. The "Global War on Terror" has meant that incidents which were previously exclusively seen in conflict areas, can occur anywhere, and clinicians who are involved in emergency trauma care may be required to manage casualties from similar terrorist attacks. An understanding of the types of devices and their pathophysiological effects is necessary to allow proper planning of mass casualty events and to allow appropriate management of the complex poly-trauma casualties they invariably cause. The aim of this review article is to firstly describe the physics and injury profile from these different devices and secondly to present the current clinical evidence that underpins their medical management.

  19. Radiation detector device for measuring ionizing radiation

    International Nuclear Information System (INIS)

    Brake, D. von der.

    1983-01-01

    The device contains a compensating filter circuit, which guarantees measurement of the radiation dose independent of the energy or independent of the energy and direction. The compensating filter circuit contains a carrier tube of a slightly absorbing metal with an order number not higher than 35, which surrounds a tubular detector and which carries several annular filter parts on its surface. (orig./HP) [de

  20. Current contact device for a superconducting magnet coil

    International Nuclear Information System (INIS)

    Hieronymus, H.

    1987-01-01

    The invention concerns a current supply device for a superconducting magnet coil to be shortcircuited, with a separating device per coil end, which contains a fixed cooled contact and a moving contact connected to a power supply device and a mechanical actuating device for closing and opening the contacts. When closing the heated contact on to the cooled contact, relatively large quantities of heat can be transferred to the cooled contact and therefore to the connected superconducting coil end and can cause normal conduction there. The invention therefore provides that the mass ratio of the cooled contact to the moving contact is at least 5:1, preferably at least 10:1, and that the cooled contact part is provided, at the end away from the contact area, with means for increasing the area, for example cooling fins and is connected to the coil end has a thermal resistance between the contact area and the coil end of at least 0.2 k/W, preferably at least 0.5 k/W per 1000 A of current to be transmitted. (orig.) [de

  1. Highly efficient red electrophosphorescent devices at high current densities

    International Nuclear Information System (INIS)

    Wu Youzhi; Zhu Wenqing; Zheng Xinyou; Sun, Runguang; Jiang Xueyin; Zhang Zhilin; Xu Shaohong

    2007-01-01

    Efficiency decrease at high current densities in red electrophosphorescent devices is drastically restrained compared with that from conventional electrophosphorescent devices by using bis(2-methyl-8-quinolinato)4-phenylphenolate aluminum (BAlq) as a hole and exciton blocker. Ir complex, bis(2-(2'-benzo[4,5-α]thienyl) pyridinato-N,C 3' ) iridium (acetyl-acetonate) is used as an emitter, maximum external quantum efficiency (QE) of 7.0% and luminance of 10000cd/m 2 are obtained. The QE is still as high as 4.1% at higher current density J=100mA/cm 2 . CIE-1931 co-ordinates are 0.672, 0.321. A carrier trapping mechanism is revealed to dominate in the process of electroluminescence

  2. Device for measuring active, reactive and apparent power

    Energy Technology Data Exchange (ETDEWEB)

    Bartosinski, E.; Wieland, J.

    1982-09-30

    The plan consists of a traditional electrodynamic mechanism for measuring power (IM) supplemented by three switches, two rectifiers, resistor, included in parallel, and phaseshifting throttle included in series with the voltage coil of the IM. This makes it possible by selection to perform three types of measurements: active power of alternating current or power of direct current, only the voltage coils and the IM current are engaged; reactive power, the resistor and the throttle are additionally engaged by the aforementioned method; complete (apparent) power--the current and the voltage are supplied directly to the IM coils, but in contrast to the first case, through rectifiers. The influence of the highest harmonic components of voltage and current which are not significant for industrial measurements can be eliminated in necessary cases using filtering devices.

  3. Research trend in thermally stimulated current method for development of materials and devices in Japan

    Science.gov (United States)

    Iwamoto, Mitsumasa; Taguchi, Dai

    2018-03-01

    Thermally stimulated current (TSC) measurement is widely used in a variety of research fields, i.e., physics, electronics, electrical engineering, chemistry, ceramics, and biology. TSC is short-circuit current that flows owing to the displacement of charges in samples during heating. TSC measurement is very simple, but TSC curves give very important information on charge behaviors. In the 1970s, TSC measurement contributed greatly to the development of electrical insulation engineering, semiconductor device technology, and so forth. Accordingly, the TSC experimental technique and its analytical method advanced. Over the past decades, many new molecules and advanced functional materials have been discovered and developed. Along with this, TSC measurement has attracted much attention in industries and academic laboratories as a way of characterizing newly discovered materials and devices. In this review, we report the latest research trend in the TSC method for the development of materials and devices in Japan.

  4. Measurement-Device-Independent Approach to Entanglement Measures

    Science.gov (United States)

    Shahandeh, Farid; Hall, Michael J. W.; Ralph, Timothy C.

    2017-04-01

    Within the context of semiquantum nonlocal games, the trust can be removed from the measurement devices in an entanglement-detection procedure. Here, we show that a similar approach can be taken to quantify the amount of entanglement. To be specific, first, we show that in this context, a small subset of semiquantum nonlocal games is necessary and sufficient for entanglement detection in the local operations and classical communication paradigm. Second, we prove that the maximum payoff for these games is a universal measure of entanglement which is convex and continuous. Third, we show that for the quantification of negative-partial-transpose entanglement, this subset can be further reduced down to a single arbitrary element. Importantly, our measure is measurement device independent by construction and operationally accessible. Finally, our approach straightforwardly extends to quantify the entanglement within any partitioning of multipartite quantum states.

  5. Polymeric salt bridges for conducting electric current in microfluidic devices

    Science.gov (United States)

    Shepodd, Timothy J [Livermore, CA; Tichenor, Mark S [San Diego, CA; Artau, Alexander [Humacao, PR

    2009-11-17

    A "cast-in-place" monolithic microporous polymer salt bridge for conducting electrical current in microfluidic devices, and methods for manufacture thereof is disclosed. Polymeric salt bridges are formed in place in capillaries or microchannels. Formulations are prepared with monomer, suitable cross-linkers, solvent, and a thermal or radiation responsive initiator. The formulation is placed in a desired location and then suitable radiation such as UV light is used to polymerize the salt bridge within a desired structural location. Embodiments are provided wherein the polymeric salt bridges have sufficient porosity to allow ionic migration without bulk flow of solvents therethrough. The salt bridges form barriers that seal against fluid pressures in excess of 5000 pounds per square inch. The salt bridges can be formulated for carriage of suitable amperage at a desired voltage, and thus microfluidic devices using such salt bridges can be specifically constructed to meet selected analytical requirements.

  6. Method and device for current driven electric energy conversion

    DEFF Research Database (Denmark)

    2012-01-01

    Device comprising an electric power converter circuit for converting electric energy. The converter circuit comprises a switch arrangement with two or more controllable electric switches connected in a switching configuration and controlled so as to provide a current drive of electric energy from...... configurations such as half bridge buck, full bridge buck, half bridge boost, or full bridge boost. A current driven conversion is advantageous for high efficient energy conversion from current sources such as solar cells or where a voltage source is connected through long cables, e.g. powerline cables for long...... an associated electric source connected to a set of input terminals. This is obtained by the two or more electric swiches being connected and controlled to short-circuit the input terminals during a part of a switching period. Further, a low pass filter with a capacitor and an inductor are provided to low pass...

  7. Observation of a current-limited double layer in a linear turbulent-heating device

    International Nuclear Information System (INIS)

    Inuzuka, H.; Torii, Y.; Nagatsu, M.; Tsukishima, T.

    1985-01-01

    Time- and space-resolved measurements of strong double layers (DLs) have been carried out for the first time on a linear turbulent-heating device, together with those of fluctuation spectra and precise current measurements. A stable stong DL is formed even when the electric current through the DL is less than the so-called Bohm value. Discussion of the formation and decay processes is given, indicating a transition from an ion-acoustic DL to a monotonic DL

  8. High current density ion beam measurement techniques

    International Nuclear Information System (INIS)

    Ko, W.C.; Sawatzky, E.

    1976-01-01

    High ion beam current measurements are difficult due to the presence of the secondary particles and beam neutralization. For long Faraday cages, true current can be obtained only by negative bias on the target and by summing the cage wall and target currents; otherwise, the beam will be greatly distorted. For short Faraday cages, a combination of small magnetic field and the negative target bias results in correct beam current. Either component alone does not give true current

  9. Measuring device for bending of beryllium reflector

    International Nuclear Information System (INIS)

    Nishida, Seiri; Sakamoto, Naoki.

    1994-01-01

    The device of the present invention can measure bending of a beryllium reflector formed in a reactor core of a nuclear reactor by a relatively easy operation. Namely, a sensor portion comprises a long-support that can be inserted to a fuel element-insertion hole disposed in the reactor and a plurality of distance sensors disposed in a longitudinal direction of the support. A supersonic wave sensor which is advantageous in the heat resistance, the size and the accuracy and can conduct measurement in water relatively easily is used as the distance sensors. However, other sensors, instead of the sensor described above, may also be used. The plurality of distance sensors detect the bending amount of the beryllium reflector in the longitudinal direction by such an easy operation of inserting such a sensor portion to the fuel element-insertion hole upon exchange of fuel elements. (I.S.)

  10. Characterization of current transport in ferroelectric polymer devices

    KAUST Repository

    Hanna, Amir

    2014-01-01

    We report the charge injection characteristics in poly(vinylidene fluoride-trifluoroethylene), P(VDF-TrFE), as a function of electrode material in metal/ferroelectric/metal device structures. Symmetric and asymmetric devices with Al, Ag, Au and Pt electrodes were fabricated to determine the dominant carrier type, injection current density, and to propose transport mechanisms in the ferroelectric polymer. Higher work function metals such as Pt are found to inject less charges compared to lower work function metals, implying n-type conduction behavior for P(VDF-TrFE) with electrons as the dominant injected carrier. Two distinct charge transport regimes were identified in the P(VDF-TrFE) devices; a Schottky-limited conduction regime for low to intermediate fields (E < 20 MV/m), and a space-charge limited conduction (SCLC) regime for high fields (20 < E < 120 MV/m). Implication of these results for degradation in P(VDF-TrFE) memory performance are discussed. © 2013 Elsevier B.V. All rights reserved.

  11. 21 CFR 864.6400 - Hematocrit measuring device.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Hematocrit measuring device. 864.6400 Section 864.6400 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) MEDICAL DEVICES HEMATOLOGY AND PATHOLOGY DEVICES Manual Hematology Devices § 864.6400 Hematocrit measuring...

  12. Burnup degree measuring device for spent fuel

    International Nuclear Information System (INIS)

    Doi, Hideo; Imaizumi, Hideki; Endo, Yasumi; Itahara, Kuniyuki.

    1994-01-01

    The present invention provides a small-sized and convenient device for measuring a burnup degree of spent fuels, which can be installed without remodelling an existent fuel storage pool. Namely, a gamma-ray detecting portion incorporates a Cd-Te detector for measuring intensity ratio of gamma-rays. A neutron detecting portion incorporates a fission counter tube. The Cd-Te detector comprises a neutron shielding member for reducing radiation damages and a background controlling plate for reducing low energy gamma-rays entering from a collimator. Since the Cd-Td detector for use in a gamma-ray spectroscopy can be used at a normal temperature and can measure even a relatively strong radiation field, it can measure the intensity of gamma-rays from Cs-137 and Cs-134 in spent fuels accurately at a resolving power of less than 10 keV. Further, in a case where a cooling period is less than one year, gamma-rays from Rh-106 and Nb-95 can also be measured. (I.S.)

  13. Current Trends in Implantable Left Ventricular Assist Devices

    Directory of Open Access Journals (Sweden)

    Jens Garbade

    2011-01-01

    Full Text Available The shortage of appropriate donor organs and the expanding pool of patients waiting for heart transplantation have led to growing interest in alternative strategies, particularly in mechanical circulatory support. Improved results and the increased applicability and durability with left ventricular assist devices (LVADs have enhanced this treatment option available for end-stage heart failure patients. Moreover, outcome with newer pumps have evolved to destination therapy for such patients. Currently, results using nonpulsatile continuous flow pumps document the evolution in outcomes following destination therapy achieved subsequent to the landmark Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure Trial (REMATCH, as well as the outcome of pulsatile designed second-generation LVADs. This review describes the currently available types of LVADs, their clinical use and outcomes, and focuses on the patient selection process.

  14. Current and Perspective Applications of Dense Plasma Focus Devices

    Science.gov (United States)

    Gribkov, V. A.

    2008-04-01

    Dense Plasma Focus (DPF) devices' applications, which are intended to support the main-stream large-scale nuclear fusion programs (NFP) from one side (both in fundamental problems of Dense Magnetized Plasma physics and in its engineering issues) as well as elaborated for an immediate use in a number of fields from the other one, are described. In the first direction such problems as self-generated magnetic fields, implosion stability of plasma shells having a high aspect ratio, etc. are important for the Inertial Confinement Fusion (ICF) programs (e.g. as NIF), whereas different problems of current disruption phenomenon, plasma turbulence, mechanisms of generation of fast particles and neutrons in magnetized plasmas are of great interest for the large devices of the Magnetic Plasma Confinement—MPC (e.g. as ITER). In a sphere of the engineering problems of NFP it is shown that in particular the radiation material sciences have DPF as a very efficient tool for radiation tests of prospect materials and for improvement of their characteristics. In the field of broad-band current applications some results obtained in the fields of radiation material sciences, radiobiology, nuclear medicine, express Neutron Activation Analysis (including a single-shot interrogation of hidden illegal objects), dynamic non-destructive quality control, X-Ray microlithography and micromachining, and micro-radiography are presented. As the examples of the potential future applications it is proposed to use DPF as a powerful high-flux neutron source to generate very powerful pulses of neutrons in the nanosecond (ns) range of its duration for innovative experiments in nuclear physics, for the goals of radiation treatment of malignant tumors, for neutron tests of materials of the first wall, blankets and NFP device's constructions (with fluences up to 1 dpa per a year term), and ns pulses of fast electrons, neutrons and hard X-Rays for brachytherapy.

  15. Current and Perspective Applications of Dense Plasma Focus Devices

    International Nuclear Information System (INIS)

    Gribkov, V. A.

    2008-01-01

    Dense Plasma Focus (DPF) devices' applications, which are intended to support the main-stream large-scale nuclear fusion programs (NFP) from one side (both in fundamental problems of Dense Magnetized Plasma physics and in its engineering issues) as well as elaborated for an immediate use in a number of fields from the other one, are described. In the first direction such problems as self-generated magnetic fields, implosion stability of plasma shells having a high aspect ratio, etc. are important for the Inertial Confinement Fusion (ICF) programs (e.g. as NIF), whereas different problems of current disruption phenomenon, plasma turbulence, mechanisms of generation of fast particles and neutrons in magnetized plasmas are of great interest for the large devices of the Magnetic Plasma Confinement--MPC (e.g. as ITER). In a sphere of the engineering problems of NFP it is shown that in particular the radiation material sciences have DPF as a very efficient tool for radiation tests of prospect materials and for improvement of their characteristics. In the field of broad-band current applications some results obtained in the fields of radiation material sciences, radiobiology, nuclear medicine, express Neutron Activation Analysis (including a single-shot interrogation of hidden illegal objects), dynamic non-destructive quality control, X-Ray microlithography and micromachining, and micro-radiography are presented. As the examples of the potential future applications it is proposed to use DPF as a powerful high-flux neutron source to generate very powerful pulses of neutrons in the nanosecond (ns) range of its duration for innovative experiments in nuclear physics, for the goals of radiation treatment of malignant tumors, for neutron tests of materials of the first wall, blankets and NFP device's constructions (with fluences up to 1 dpa per a year term), and ns pulses of fast electrons, neutrons and hard X-Rays for brachytherapy

  16. Measuring Electrical Current: The Roads Not Taken

    Science.gov (United States)

    Greenslade, Thomas B., Jr.

    2011-01-01

    Recently I wrote about the standard Weston meter movement, that is at the heart of all modern analogue current measurements. Now I will discuss other techniques used to measure electric current that, despite being based on valid physical principles, are largely lost in technological history.

  17. Marine current energy devices: Current status and possible future applications in Ireland

    International Nuclear Information System (INIS)

    Rourke, Fergal O.; Boyle, Fergal; Reynolds, Anthony

    2010-01-01

    There is a growing demand for the use of renewable energy technologies to generate electricity due to concerns over climate change. The oceans provide a huge potential resource of energy. Energy extraction using marine current energy devices (MCEDs) offers a sustainable alternative to conventional sources and a predictable alternative to other renewable energy technologies. A MCED utilises the kinetic energy of the tides as opposed to the potential energy which is utilised by a tidal barrage. Over the past decade MCEDs have become an increasingly popular method of energy extraction. However, marine current energy technology is still not economically viable on a large scale due to its current stage of development. Ireland has an excellent marine current energy resource as it is an island nation and experiences excellent marine current flows. This paper reviews marine current energy devices, including a detailed up-to-date description of the current status of development. Issues such as network integration, economics, and environmental implications are addressed as well as the application and costs of MCEDs in Ireland. (author)

  18. Fast measure proceeding of weak currents

    International Nuclear Information System (INIS)

    Taieb, J.

    1953-01-01

    The process of fast measure of the weak currents that we are going to describe briefly apply worthy of the provided currents by the sources to elevated value internal resistance, as it is the case for the ionization chamber, the photocells, mass spectroscopic tubes. The problem to measure weak currents is essentially a problem of amplifier and of input circuit. We intended to achieve a whole amplifier and input circuit with advanced performances, meaning that for a measured celerity we wanted to have an signal/noise ratio the most important as in the classic systems and for a same report signal/noise a more quickly done measure. (M.B.) [fr

  19. A Practical Device for Measuring the Luminance Distribution

    Directory of Open Access Journals (Sweden)

    Thijs Kruisselbrink

    2017-06-01

    Full Text Available Various applications in building lighting such as automated daylight systems, dynamic lighting control systems, lighting simulations, and glare analyzes can be optimized using information on the actual luminance distributions of the surroundings. Currently, commercially available luminance distribution measurement devices are often not suitable for these kind of applications or simply too expensive for broad application. This paper describes the development of a practical and autonomous luminance distribution measurement device based on a credit card-sized single-board computer and a camera system. The luminance distribution was determined by capturing High Dynamic Range images and translating the RGB information to the CIE XYZ color space. The High Dynamic Range technology was essential to accurately capture the data needed to calculate the luminance distribution because it allows to capture luminance ranges occurring in real scenarios. The measurement results were represented in accordance with established methods in the field of daylighting. Measurements showed that the accuracy of the luminance distribution measurement device ranged from 5% to 20% (worst case which was deemed acceptable for practical measurements and broad applications in the building realm.

  20. 32 CFR 634.27 - Speed-measuring devices.

    Science.gov (United States)

    2010-07-01

    ... 32 National Defense 4 2010-07-01 2010-07-01 true Speed-measuring devices. 634.27 Section 634.27 National Defense Department of Defense (Continued) DEPARTMENT OF THE ARMY (CONTINUED) LAW ENFORCEMENT AND CRIMINAL INVESTIGATIONS MOTOR VEHICLE TRAFFIC SUPERVISION Traffic Supervision § 634.27 Speed-measuring devices. Speed-measuring devices will be...

  1. Local eddy current measurements in pulsed fields

    Energy Technology Data Exchange (ETDEWEB)

    Espina-Hernandez, J.H. [SEPI-Electronica, ESIME-IPN, UPALM Edif. ' Z' . Zacatenco, Mexico DF 07738 (Mexico)], E-mail: jhespina@gmail.com; Groessinger, R. [Institute of Solid State Physics, Vienna University of Technology, Wiedner Hauptstrasse 8-10, A-1040 Vienna (Austria); Hallen, J.M. [Departamento de Ingenieria Metalurgica, IPN-ESIQIE, UPALM Edif. 7, Zacatenco, Mexico DF 07738 (Mexico)

    2008-07-15

    This work presents new eddy current measurements in pulsed fields. A commercial point pick-up coil is used to detect the induction signal along the radius of Cu and Al samples with cylindrical shape and diameters between 5 and 35 mm. Local eddy current measurements were performed on the surface of conducting materials due to the small dimensions of the coil. A simple electrical circuit, used as a model, is proposed to describe the local eddy current effect in pulsed fields. The proposed model allows to calculate the phase shift angle between the signal proportional to eddy currents and the applied external field in a pulsed field magnetometer.

  2. Zero current measurements using the Rogowski coil

    International Nuclear Information System (INIS)

    Gregor, J.; Jakubova, I.; Kadlec, P.; Senk, J.; Vavra, Z.

    1997-01-01

    The zero current measurements using the Rogowski coil carried out on the model of the extinguishing chamber of hv SF 6 circuit breaker with self-flow generation are presented in the paper. The time course of the post-arc current obtained by numerical integration of the measured induced voltage of the Rogowski coil gives information not only about the value of the residual current after the successful interruption but also about the current changes connected with the dynamic behaviour of the arc during its quenching. (author)

  3. 21 CFR 864.5950 - Blood volume measuring device.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Blood volume measuring device. 864.5950 Section 864.5950 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) MEDICAL DEVICES HEMATOLOGY AND PATHOLOGY DEVICES Automated and Semi-Automated Hematology Devices...

  4. Measurement-Device-Independent Quantum Cryptography

    Science.gov (United States)

    Tang, Zhiyuan

    Quantum key distribution (QKD) enables two legitimate parties to share a secret key even in the presence of an eavesdropper. The unconditional security of QKD is based on the fundamental laws of quantum physics. Original security proofs of QKD are based on a few assumptions, e.g., perfect single photon sources and perfect single-photon detectors. However, practical implementations of QKD systems do not fully comply with such assumptions due to technical limitations. The gap between theory and implementations leads to security loopholes in most QKD systems, and several attacks have been launched on sophisticated QKD systems. Particularly, the detectors have been found to be the most vulnerable part of QKD. Much effort has been put to build side-channel-free QKD systems. Solutions such as security patches and device-independent QKD have been proposed. However, the former are normally ad-hoc, and cannot close unidentified loopholes. The latter, while having the advantages of removing all assumptions on devices, is impractical to implement today. Measurement-device-independent QKD (MDI-QKD) turns out to be a promising solution to the security problem of QKD. In MDI-QKD, all security loopholes, including those yet-to-be discovered, have been removed from the detectors, the most critical part in QKD. In this thesis, we investigate issues related to the practical implementation and security of MDI-QKD. We first present a demonstration of polarization-encoding MDI-QKD. Taking finite key effect into account, we achieve a secret key rate of 0.005 bit per second (bps) over 10 km spooled telecom fiber, and a 1600-bit key is distributed. This work, together with other demonstrations, shows the practicality of MDI-QKD. Next we investigate a critical assumption of MDI-QKD: perfect state preparation. We apply the loss-tolerant QKD protocol and adapt it to MDI-QKD to quantify information leakage due to imperfect state preparation. We then present an experimental demonstration of

  5. Current measurement system utilizing cryogenic techniques for the absolute measurement of the magnetic flux quantum

    International Nuclear Information System (INIS)

    Endo, T.; Murayama, Y.; Sakamoto, Y.; Sakuraba, T.; Shiota, F.

    1989-01-01

    A series of systems composed of cryogenic devices such as a Josephson potentiometer and a cryogenic current comparator has been proposed and developed to precisely measure a current with any value up to 1 A. These systems will be used to measure the injected electrical energy with an uncertainty of the order of 0.01 ppm or less in the absolute measurement of the magnetic flux quantum by superconducting magnetic levitation. Some preliminary experiments are described

  6. Continuous measurement of an atomic current

    Science.gov (United States)

    Laflamme, C.; Yang, D.; Zoller, P.

    2017-04-01

    We are interested in dynamics of quantum many-body systems under continuous observation, and its physical realizations involving cold atoms in lattices. In the present work we focus on continuous measurement of atomic currents in lattice models, including the Hubbard model. We describe a Cavity QED setup, where measurement of a homodyne current provides a faithful representation of the atomic current as a function of time. We employ the quantum optical description in terms of a diffusive stochastic Schrödinger equation to follow the time evolution of the atomic system conditional to observing a given homodyne current trajectory, thus accounting for the competition between the Hamiltonian evolution and measurement back action. As an illustration, we discuss minimal models of atomic dynamics and continuous current measurement on rings with synthetic gauge fields, involving both real space and synthetic dimension lattices (represented by internal atomic states). Finally, by "not reading" the current measurements the time evolution of the atomic system is governed by a master equation, where—depending on the microscopic details of our CQED setups—we effectively engineer a current coupling of our system to a quantum reservoir. This provides interesting scenarios of dissipative dynamics generating "dark" pure quantum many-body states.

  7. A Simple Device For Measuring Skin Friction

    Directory of Open Access Journals (Sweden)

    Gupta A.B

    1995-01-01

    Full Text Available A simple device for measuring skin friction in vivo is described. The frictional coefficient of normal Indian skin and the effect of hydration and application of talc and glycerol on the frictional coefficient and also the friction of ichthyotic skin have been determined with its help. The average value of friction of friction of normal India skin at forearm is found to be 0.41 +- 0.08, the hydration raises the value to 0.71 +- 0.11 and the effect of glycerol is also to school it up to 0.70+- 0.05, almost equal to that of water. The effect of talc however is opposite and its application lowers the friction to 0.21+-0.07. The mean coeff of friction for ichthyotic skin is found to be 0.21+- 0.0.5, which closely agrees with talc-treated normal skin. A good positive correlation (p<0.01 between friction and sebum level at skin site, with r = 0.64, has been observed.

  8. Cross-Device Tracking: Measurement and Disclosures

    Directory of Open Access Journals (Sweden)

    Brookman Justin

    2017-04-01

    Full Text Available Internet advertising and analytics technology companies are increasingly trying to find ways to link behavior across the various devices consumers own. This cross-device tracking can provide a more complete view into a consumer’s behavior and can be valuable for a range of purposes, including ad targeting, research, and conversion attribution. However, consumers may not be aware of how and how often their behavior is tracked across different devices. We designed this study to try to assess what information about cross-device tracking (including data flows and policy disclosures is observable from the perspective of the end user. Our paper demonstrates how data that is routinely collected and shared online could be used by online third parties to track consumers across devices.

  9. The insertion device magnetic measurement facility: Prototype and operational procedures

    International Nuclear Information System (INIS)

    Burkel, L.; Dejus, R.; Maines, J.; O'Brien, J.; Vasserman, I.; Pfleuger, J.

    1993-03-01

    This report is a description of the current status of the magnetic measurement facility and is a basic instructional manual for the operation of the facility and its components. Please refer to the appendices for more detailed information about specific components and procedures. The purpose of the magnetic measurement facility is to take accurate measurements of the magnetic field in the gay of the IDs in order to determine the effect of the ID on the stored particle beam and the emitted radiation. The facility will also play an important role when evaluating new ideas, novel devices, and inhouse prototypes as part of the ongoing research and development program at the APS. The measurements will be performed with both moving search coils and moving Hall probes. The IDs will be evaluated by computer modeling of the emitted radiation for any given (measured) magnetic field map. The quality of the magnetic field will be described in terms of integrated multipoles for the effect on Storage Ring performance and in terms of the derived trajectories for the emitted radiation. Before being installed on the Storage Ring, every device will be measured and characterized to assure that it is compatible with Storage Ring requirements and radiation specifications. The accuracy that the APS needs to achieve for magnetic measurements will be based on these specifications

  10. Guidelines on calibration of neutron measuring devices

    International Nuclear Information System (INIS)

    Burger, G.

    1988-01-01

    The International Atomic Energy Agency and the World Health Organization have agreed to establish an IAEA/WHO Network of Secondary Standard Dosimetry Laboratories (SSDLs) in order to improve accuracy in applied radiation dosimetry throughout the world. These SSDLs must be equipped with, and maintain, secondary standard instruments, which have been calibrated against primary standards, and must be nominated by their governments for membership of the network. The majority of the existing SSDLs were established primarily to work with photon radiation (X-rays and gamma rays). Neutron sources are, however, increasingly being applied in industrial processes, research, nuclear power development and radiation biology and medicine. Thus, it is desirable that the SSDLs in countries using neutron sources on a regular basis should also fulfil the minimum requirements to calibrate neutron measuring devices. It is the primary purpose of this handbook to provide guidance on calibration of instruments for radiation protection. A calibration laboratory should also be in a position to calibrate instrumentation being used for the measurement of kerma and absorbed dose and their corresponding rates. This calibration is generally done with photons. In addition, since each neutron field is usually contaminated by photons produced in the source or by scatter in the surrounding media, neutron protection instrumentation has to be tested with respect to its intrinsic photon response. The laboratory will therefore need to possess equipment for photon calibration. This publication deals primarily with methods of applying radioactive neutron sources for calibration of instrumentation, and gives an indication of the space, manpower and facilities needed to fulfil the minimum requirements of a calibration laboratory for neutron work. It is intended to serve as a guide for centres about to start on neutron dosimetry standardization and calibration. 94 refs, 8 figs, 12 tabs

  11. Advanced devices and systems for radiation measurements

    International Nuclear Information System (INIS)

    Knoll, G.F.; Wehe, D.K.; He, Z.; Barrett, C.; Miyamoto, J.

    1996-06-01

    The authors' most recent work continues their long-standing efforts to develop semiconductor detectors based on the collection of only a single type of charge carrier. Their best results are an extension of the principle of coplanar electrodes first described by Paul Luke of Lawrence Berkeley Laboratory 18 months ago. This technique, described in past progress reports, has the effect of deriving an output signal from detectors that depends only on the motion of carriers close to one surface. Since nearly all of these carriers are of one type (electrons) that are attracted to that electrode, the net effect is to nearly eliminate the influence of hole motion on the properties of the output signal. The result is that the much better mobility of electrons in compound semiconductors materials such as CZT can now be exploited without the concurrent penalty of poor hole collection. They have also developed new techniques in conjunction with the coplanar electrode principle that extends the technique into a new dimension. By proper processing of signals from the opposite electrode (the cathode) from the coplanar surface, they are able to derive a signal that is a good indication of the depth of interaction at which the charge carriers were initially formed. They have been the first group to demonstrate this technique, and examples of separate pulse height spectra recorded at a variety of different depths of interaction are shown in several of the figures that follow. Obtaining depth information is one step in the direction of obtaining volumetric point-of-interaction information from the detector. If one could known the coordinates of each specific interaction, then corrections could be applied to account for the inhomogeneities that currently plague many room-temperature devices

  12. Device for measuring the temperature of flowing hot gases

    Energy Technology Data Exchange (ETDEWEB)

    Reed, R D

    1977-05-12

    The invention pertains to a device to measure the temperature of a hot gas flowing through a closed tube. The device will have a simple and inexpensive design and avoid heat losses due to heat radiation near the thermal sensor.

  13. Automatic system for ionization chamber current measurements

    International Nuclear Information System (INIS)

    Brancaccio, Franco; Dias, Mauro S.; Koskinas, Marina F.

    2004-01-01

    The present work describes an automatic system developed for current integration measurements at the Laboratorio de Metrologia Nuclear of Instituto de Pesquisas Energeticas e Nucleares. This system includes software (graphic user interface and control) and a module connected to a microcomputer, by means of a commercial data acquisition card. Measurements were performed in order to check the performance and for validating the proposed design

  14. Measurement of toroidal plasma current in RF heated helical plasmas

    International Nuclear Information System (INIS)

    Besshou, Sakae

    1993-01-01

    This report describes the measurement of toroidal plasma current by a semiflexible Rogowski coil in a helical vacuum chamber. A Rogowski coil measures the toroidal plasma current with a resolution of 0.1 kA, frequency range of up to 1 kHz and sensitivity of 6.5 x 10 -9 V · s/A. We measured the spontaneous toroidal plasma current (from -1.2 to +1.2 kA) under electron cyclotron resonance heating at 0.94 T toroidal field in the Heliotron-E device. We found that the measured direction of toroidal plasma current changes its sign as in the predicted behavior of a neoclassical diffusion-driven bootstrap current, depending on the horizontal position of the plasma column. We explain the observed plasma currents in terms of the compound phenomenon of an ohmic current and a neoclassical diffusion-driven current. The magnitude of the neoclassical current component is smaller than the value predicted by a collisionless neoclassical theory. (author)

  15. Method and device for measuring the smoke concentration in air

    International Nuclear Information System (INIS)

    Rennemo, B.

    1994-01-01

    The patent deals with a method and a device for measuring the smoke concentration in air. In a smoke chamber are located two electrodes, connected to a voltage source for forming a circuit in which a DC current flows. A radioactive radiation source to ionize the air molecules is located in the vicinity of the smoke chamber, so that the number of ionized air molecules which are formed is dependent upon the radiation intensity of the ion source and the concentration of smoke particles in the smoke chamber. The charging voltage will further imply that a cloud of high ion concentration is built up close to the surface of the electrodes. The ion cloud will be discharged capacitively upon a plurality of short voltages pulses applied to the electrodes to thereby result in current pulses substantially greater than the DC current flowing through the chamber. 8 figs

  16. Advances in superconductivity: new materials, critical currents and devices

    International Nuclear Information System (INIS)

    Pinto, R.; Malik, S.K.; Grover, A.K.; Ayyub, P.

    1997-01-01

    The discovery of superconductivity in the cuprates produced an explosive growth in research, driven by the quest for higher and higher superconducting transition temperatures. In the initial stages, the excitement was tremendous both in the physical sciences and in engineering. However, the complexity of the new materials on the one hand, and the absence of a viable theory on the other, have made further developments much more difficult. It is to be expected therefore, that the early excitement and the subsequent rapid advances have paved the way for more systematic and detailed studies of all aspects of superconductivity. The International Symposium was intended to provide a forum to review the progress in selected areas in superconductivity. The emphasis was on experimental and theoretical studies of the new superconductors, advances in the theoretical understanding, progress in studies of flux pinning and vortex dynamics which affect critical currents, and developments of novel material synthesis methods. Recent developments in the twin areas of thin films and devices were extensively discussed during the symposium. Papers relevant to INIS are indexed separately

  17. Gate tunneling current and quantum capacitance in metal-oxide-semiconductor devices with graphene gate electrodes

    Science.gov (United States)

    An, Yanbin; Shekhawat, Aniruddh; Behnam, Ashkan; Pop, Eric; Ural, Ant

    2016-11-01

    Metal-oxide-semiconductor (MOS) devices with graphene as the metal gate electrode, silicon dioxide with thicknesses ranging from 5 to 20 nm as the dielectric, and p-type silicon as the semiconductor are fabricated and characterized. It is found that Fowler-Nordheim (F-N) tunneling dominates the gate tunneling current in these devices for oxide thicknesses of 10 nm and larger, whereas for devices with 5 nm oxide, direct tunneling starts to play a role in determining the total gate current. Furthermore, the temperature dependences of the F-N tunneling current for the 10 nm devices are characterized in the temperature range 77-300 K. The F-N coefficients and the effective tunneling barrier height are extracted as a function of temperature. It is found that the effective barrier height decreases with increasing temperature, which is in agreement with the results previously reported for conventional MOS devices with polysilicon or metal gate electrodes. In addition, high frequency capacitance-voltage measurements of these MOS devices are performed, which depict a local capacitance minimum under accumulation for thin oxides. By analyzing the data using numerical calculations based on the modified density of states of graphene in the presence of charged impurities, it is shown that this local minimum is due to the contribution of the quantum capacitance of graphene. Finally, the workfunction of the graphene gate electrode is extracted by determining the flat-band voltage as a function of oxide thickness. These results show that graphene is a promising candidate as the gate electrode in metal-oxide-semiconductor devices.

  18. Cryogenic current comparators for precise ion beam current measurements

    International Nuclear Information System (INIS)

    Kurian, Febin

    2015-01-01

    The planned Facility for Antiproton and Ion Research (FAIR) at GSI has to cope with a wide range of beam intensities in its high-energy beam transport systems and in the storage rings. To meet the requirements of a non-intercepting intensity measurement down to nA range, it is planned to install a number of Cryogenic Current Comparator (CCC) units at different locations in the FAIR beamlines. In this work, the first CCC system for intensity measurement of heavy ion beams, which was developed at GSI, was re-commissioned and upgraded to be used as a 'GSI - CCC prototype' for extensive optimization and development of an improved CCC for FAIR. After installation of a new SQUID sensor and related electronics, as well as implementation of improved data acquisition components, successful beam current measurements were performed at a SIS18 extraction line. The measured intensity values were compared with those of a Secondary Electron Monitor (SEM). Furthermore, the spill-structure of a slowly extracted beam was measured and analyzed, investigating its improvement due to bunching during the slow-extraction process. Due to the extreme sensitivity of the superconducting sensor, the determined intensity values as well as the adjustment of the system for optimal performance are strongly influenced by the numerous noise sources of the accelerators environment. For this reason, detailed studies of different effects caused by noise have been carried out, which are presented together with proposals to reduce them. Similarly, studies were performed to increase the dynamic range and overcome slew rate limitations, the results of which are illustrated and discussed as well. By combining the various optimizations and characterizations of the GSI CCC prototype with the experiences made during beam operation, criteria for a more efficient CCC System could be worked out, which are presented in this work. The details of this new design are worked out with respect to the

  19. Mechanical Design of the Intensity Measurement Devices for the LHC

    CERN Document Server

    Belorhad, D; Odier, P; Thoulet, S

    2008-01-01

    The intensity measurement for the LHC ring is provided by eight current transformers (2×DCCT, 2×FBCT per beam). The measurement resolution of 1?Arms at 1s average for the DCCTs and ±10^9p in 25ns for the FBCTs is required. Such constraints call for low noise electronics and a compact magnetically shielded mechanical design. Correct integration of these devices into the vacuum system also requires the vacuum chambers equipped with the non-evaporable getter (NEG) film. The NEG is activated by heating the vacuum chamber to 200?C and more. Such temperatures affect the structure of the magnetic materials, which form the base part of the intensity measurement devices, and degrade their performace. A cooling circuit is needed. Due to the mechanical constraints, the cooling circuit, as well as heating element must form an integral part of the design. The aim of this paper is to present the solutions to these problems and discuss the mechanical construction of the DCCTs and FBCTs currently being installed in the LH...

  20. Confinement studies of a high current density RFP in the Extrap T1 Upgrade device

    International Nuclear Information System (INIS)

    Drake, J.R.; Brzozowski, J.H.; Brunsell, P.; Hellblom, G.; Karlsson, P.; Mazur, S.; Nordlund, P.; Welander, A.; Zastrow, K.D.

    1992-01-01

    Confinement studies have been carried out on the Extrap T1 device operated in the reversed field pinch (RFP) mode. Extrap T1 is a small device with a major radius of R=0.5 m and a high aspect ratio, R/a=8.9. For these experiments, the device has been operated with a resistive shell with measured, toroidally-averaged flux penetration times of τ sv = 500μs (vertical) and τ sR =300μs (radial). The pulse lengths are about 600 μs, which is slightly longer than the shell penetration time. The purpose of these experiments is to study energy confinement in a high aspect-ratio, high current-density RFP device with a resistive shell. The device can be operated with high current densities which exceed 20 MAm -2 on axis. For these discharges, the average electron density is relatively high, ≅ 1x10 20 m -3 . Therefore, although the average current density exceeds 5 MAm -2 , the important parameter / ≅ I/N is maintained less than 1x10 -13 Am, where N is the line density. The plasma diagnostics for the device include a single chord CO 2 laser interferometer ( ), single point Thomson scattering (T e , n o ), VUV and visible spectroscopy (T e , Z eff ) surface barrier diodes for soft X-ray measurements (T e ), bolometry (P rad ), surface probes (Γ p ,T i ) and comprehensive magnetic diagnostics for both equilibrium and magnetic fluctuation studies. (author) 5 refs., 1 fig., 1 tab

  1. Eddy currents in pulsed field measurements

    International Nuclear Information System (INIS)

    Kuepferling, M.; Groessinger, R.; Wimmer, A.; Taraba, M.; Scholz, W.

    2002-01-01

    Full text: One problem of pulsed field magnetometry is an error in magnetization, which appears in measurements of conducting samples. This error is due to eddy currents induced by a time varying field. To allow predictions how eddy currents exert influence on the hysteresis loop, systematic experimental and theoretical studies of pulsed field measurements of metallic samples were performed. The theoretical studies include analytical calculations as well as numerical ones using a 2D finite element software. In the measurements three physical parameters have been varied: i) the conductivity of the sample by using two different materials, in this case technical Cu and Al ii) size and shape of the sample by using cylinders, spheres and cuboids iii) the pulse duration of the external field by changing the capacitor battery from 8mF ( =9.1ms) to 24mF ( =15.7ms). The time dependence of the external field corresponds with a pulsed damped harmonic oscillation with a maximum value of 5.2T. The samples were studied in the as cast state (after machining) as well as after heat treatment. Theoretical calculations showed not only good agreement with the absolute values of the measured eddy current m agnetization , they also gave an explanation of the shape of the eddy current hysteresis and the dependence of the eddy current 'magnetization' on parameters as pulse duration of the external field and conductivity of the sample. (author)

  2. Digital Thickness Measurement of a Transparent Plastic Orthodontic Device

    Science.gov (United States)

    Kim, Yoon-Hwan; Rhim, Sung-Han

    2018-05-01

    A transparent orthodontic device is used to move the teeth to the final calibration position to form a proper set of teeth. Because the uniform thickness of the device plays an important role in tooth positioning, the accuracy of the device's thickness profile is important for effective orthodontic treatment. However, due to the complexity of the device's geometry and the transparency of the device's material, measuring the complete thickness profile has been difficult. In the present study, a new optical scanning method to measure the thickness profile of transparent plastic orthodontic devices is proposed and evaluated by using scanning electron microscopy (SEM). The error of the new measurement method is less than ±18 μm. The new method can be used to measure the thickness of non-specific, multi-curved, transparent orthodontic devices.

  3. Development of an automatized device for eddy current testing

    International Nuclear Information System (INIS)

    David, B.; Benoist, B.; Champonnois, F.; Viard, J.

    1990-01-01

    To improve performances in eddy current studies, several tools were developed in the Saclay Laboratory: - ELEFANT is a very flexible software which allows a rapid creation of new NDT applications, preserving anytime a full compatibility between computers, applications, data... used under the same ELEFANT controlkey. - Modulo'STA is a modular hardware designed to be the flexible interface between the probe and the computer, driving the measurement, the probe movement... - rotating EC probes with local rotation facilities allow the control of very long steam generator tubes, even bent [fr

  4. Leakage current measurement in transformerless PV inverters

    DEFF Research Database (Denmark)

    Kerekes, Tamas; Sera, Dezso; Mathe, Laszlo

    2012-01-01

    Photovoltaic (PV) installations have seen a huge increase during the last couple of years. Transformerless PV inverters are gaining more share of the total inverter market, due to their high conversion efficiency, small weight and size. Nevertheless safety should have an important role in case...... of these tranformerless systems, due to the missing galvanic isolation. Leakage and fault current measurement is a key issue for these inverter topologies to be able to comply with the required safety standards. This article presents the test results of two different current measurement sensors that were suggested...

  5. Performance and Pain Tolerability of Current Diagnostic Allergy Skin Prick Test Devices.

    Science.gov (United States)

    Tversky, Jody R; Chelladurai, Yohalakshmi; McGready, John; Hamilton, Robert G

    2015-01-01

    Allergen skin prick testing remains an essential tool for diagnosing atopic disease and guiding treatment. Sensitivity needs to be defined for newly introduced devices. Our aim was to compare the performance of 10 current allergy skin prick test devices. Single- and multiheaded skin test devices (n = 10) were applied by a single operator in a prospective randomized manner. Histamine (1 and 6 mg/mL) and control diluent were introduced at 6 randomized locations onto the upper and lower arms of healthy subjects. Wheal and flare reactions were measured independently by 2 masked technicians. Twenty-four subjects provided consent, and 768 skin tests were placed. Mean wheal diameter among devices differed from 3.0 mm (ComforTen; Hollister-Stier, Spokane, Wash) to 6.8 mm (UniTest PC; Lincoln Diagnostics, Decatur, Ill) using 1 mg/mL histamine (P Diagnostics, Decatur, Ill; and Sharp-Test; Panatrex, Placentia, Calif) using 6 mg/mL histamine (P pain score of less than 4 on a 10-point visual analog scale. Pain scores were higher among women, but this did not reach statistical significance. The Multi-Test PC and the UniTest PC had the lowest pain scores compared with the other devices. All 10 skin prick test devices displayed good analytical sensitivity and specificity; however, 3 mm cannot arbitrarily be used as a positive threshold. The use of histamine at 1 mg/mL is unacceptable for certain devices but may be preferable for the most sensitive devices. On average, there was no pain score difference between multiheaded and single-head devices. Copyright © 2015 American Academy of Allergy, Asthma & Immunology. Published by Elsevier Inc. All rights reserved.

  6. Low Current Surface Flashover for Initiation of Electric Propulsion Devices

    Science.gov (United States)

    Dary, Omar G.

    new LCSF assembly (flashover current was limited to reliably ignite high-current arcs (10A-12A arc were used in this work). Ignition of the high-current arc was observed at three different cases of LCSF with limiting currents 100 mA, 33 mA and 20 mA respectively. Plasma parameter measurements were conducted with variety of Langmuir probes inside the LCSF plume. Ion currents created by the LCSF were primarily expelled directly perpendicular from the insulator surface. The plasma expansion for the LCSF assembly was measured to be 2 x 106-6 x 106 cm/s. Plasma density was measured to range 10 10-1011 cm-3. The plasma density was maximal near the LCSF assembly and quickly reduced radially. Temporal decay of the plasma was observed on a time scale of about 5 micros after the LCSF event. The results of this work are significant for creation of ignitor for micropropulsion systems. LCSF system offers reliable triggering for numerous ignition pulses for entire lifetime of the micropropulsion system and reduces complexity and volume of the system by excluding moving parts and the need for an external gas tanks.

  7. In-vitro diagnostic devices introduction to current point-of-care diagnostic devices

    CERN Document Server

    Cheng, Chao-Min; Chen, Chien-Fu

    2016-01-01

    Addressing the origin, current status, and future development of point-of-care diagnostics, and serving to integrate knowledge and tools from Analytical Chemistry, Bioengineering, Biomaterials, and Nanotechnology, this book focusses on addressing the collective and combined needs of industry and academia (including medical schools) to effectively conduct interdisciplinary research. In addition to summarizing and detailing developed diagnostic devices, this book will attempt to point out the possible future trends of development for point-of-care diagnostics using both scientifically based research and practical engineering needs with the aim to help novices comprehensively understand the development of point-of-care diagnostics. This includes demonstrating several common but critical principles and mechanisms used in point-of-care diagnostics that address practical needs (e.g., disease or healthcare monitoring) using two well-developed examples so far: 1) blood glucose meters (via electrochemistry); and, 2) p...

  8. Technique and Calculation Results of Currents and Voltages in the Circuits of the Measuring Element of the Protection Device of the Transmission Line Based on the Control of Transient Processes

    Science.gov (United States)

    Lachugin, V. F.; Kulikov, A. L.; Platonov, P. S.; Vucolov, V. Yu.

    2017-12-01

    The specifics of generation of the signals of current and voltage in the circuits of a directional element of wave relay protection during short circuit (SC) in overhead power transmission lines are considered. The computing method of transient processes in the protection circuits, including frequency filters, that attenuate the parameters of currents and voltages of the mode taking into account the higher harmonic components and probable deviations of the frequency of transmission line from the rated value is presented. It is revealed that it is advisable to implement the measuring circuits of the directional elements of wave relay protection with the three-section filter attenuating the frequencies from 45 to 55 Hz and the low pass filter with cutoff frequency that does not exceed 1 kHz.

  9. Current perspectives in percutaneous atrial septal defect closure devices

    Directory of Open Access Journals (Sweden)

    Bissessor N

    2015-07-01

    Full Text Available N Bissessor1–4 1Department of Cardiology, The Epworth Hospital, Melbourne, VIC, Australia; 2Division of Interventional Cardiology, The Alfred Hospital, Melbourne, VIC, Australia; 3Department of Clinical Science, Charles Sturt University Albury Campus, NSW, Australia; 4Heart Foundation, Griffith University, QLD, Australia Abstract: In the last decade, percutaneous atrial septal defect (ASD closure has become the treatment of choice in most clinical presentations of ASD. Percutaneous ASD closure has established procedural safety through operator experience and improved device structure and deliverability. There have also been advances in diagnostic capabilities. Devices have evolved from large bulky meshes to repositionable, minimal residual mesh content that easily endothelializes and conforms well to surrounding structures. Biodegradable technology has been introduced and will be closely watched as a future option. The evolution of ASD closure device usage in the last four decades incorporates development that minimizes a wide range of serious side effects that have been reported over the years. Complications reported in the literature include thrombus formation, air embolization, device embolization, erosions, residual shunts, and nickel hypersensitivity. Modern devices have intermediate to long term data with outcomes that have been favorable. Devices are available in multiple sizes with improved delivery mechanisms to recapture, reposition, and safely close simple and complex ASDs amenable to percutaneous closure. In this review, commonly used devices and deployment procedures are discussed together with a look at devices that show promise for the future. Keywords: ASD, congenital, Amplatzer, Gore Helex, Biostar, Figulla

  10. Characterization of current transport in ferroelectric polymer devices

    KAUST Repository

    Hanna, Amir; Bhansali, Unnat Sampatraj; Khan, Yasser; Alshareef, Husam N.

    2014-01-01

    We report the charge injection characteristics in poly(vinylidene fluoride-trifluoroethylene), P(VDF-TrFE), as a function of electrode material in metal/ferroelectric/metal device structures. Symmetric and asymmetric devices with Al, Ag, Au and Pt

  11. Pressure and Relative Humidity Measurement Devices for Mars 2020 Rover

    Science.gov (United States)

    Hieta, M.; Genzer, M.; Nikkanen, T.; Haukka, H.; Harri, A.-M.; Polkko, J.; Rodriguez-Manfredi, J. A.

    2017-09-01

    One of the scientific payloads onboard the NASA Mars 2020 rover mission is Mars Environmental Dynamic Analyzer (MEDA): a set of environmental sensors for Mars surface weather measurements. Finnish Meteorological Institute (FMI) provides a pressure measurement device (MEDA PS) and a relative humidity measurement device (MEDA HS) for MEDA.

  12. Control device intended for a gamma radiation measuring instrument

    International Nuclear Information System (INIS)

    1976-01-01

    This invention concerns a monitoring device for a gamma radiation measuring instrument or radiation meter, in which the radiation to be measured brings about, inter alia, the ionisation of a gas and the generation of current pulses. The dial of this meter is generally calibrated in roentgens per hour, i.e. in radiation rate units. This instrument of very simple design is remarkable for its operating reliability. Preferably placed at the inlet to a radioactive area, it enables every user of a ratemeter to check, over the entire measuring range of this instrument, its proper operation prior to entering the area. To this effect, the monitoring device in question has a thick wall lead castle, having an internal cavity in which is mounted a radioactive source delivering a gamma radiation with given constant characteristics, through a measurement window closed by a calibrated plug. Lead doors articulated on the castle can be superimposed on this window to bring about a given attenuation of the radiation coming from the source and delivered to the exterior of the castle [fr

  13. Current challenges for clinical trials of cardiovascular medical devices.

    Science.gov (United States)

    Zannad, Faiez; Stough, Wendy Gattis; Piña, Ileana L; Mehran, Roxana; Abraham, William T; Anker, Stefan D; De Ferrari, Gaetano M; Farb, Andrew; Geller, Nancy L; Kieval, Robert S; Linde, Cecilia; Redberg, Rita F; Stein, Kenneth; Vincent, Alphons; Woehrle, Holger; Pocock, Stuart J

    2014-07-15

    Several features of cardiovascular devices raise considerations for clinical trial conduct. Prospective, randomized, controlled trials remain the highest quality evidence for safety and effectiveness assessments, but, for instance, blinding may be challenging. In order to avoid bias and not confound data interpretation, the use of objective endpoints and blinding patients, study staff, core labs, and clinical endpoint committees to treatment assignment are helpful approaches. Anticipation of potential bias should be considered and planned for prospectively in a cardiovascular device trial. Prospective, single-arm studies (often referred to as registry studies) can provide additional data in some cases. They are subject to selection bias even when carefully designed; thus, they are generally not acceptable as the sole basis for pre-market approval of high risk cardiovascular devices. However, they complement the evidence base and fill the gaps unanswered by randomized trials. Registry studies present device safety and effectiveness in day-to-day clinical practice settings and detect rare adverse events in the post-market period. No single research design will be appropriate for every cardiovascular device or target patient population. The type of trial, appropriate control group, and optimal length of follow-up will depend on the specific device, its potential clinical benefits, the target patient population and the existence (or lack) of effective therapies, and its anticipated risks. Continued efforts on the part of investigators, the device industry, and government regulators are needed to reach the optimal approach for evaluating the safety and performance of innovative devices for the treatment of cardiovascular disease. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  14. Current status of the regulation for medical devices

    OpenAIRE

    Shah Anuja; Goyal R

    2008-01-01

    In the light of escalating use of medical devices, stringent regulatory standards are required to ensure that the devices are safe, well studied and have least adverse reactions. Recently introduced guidelines and the amendment in the law will provide adequate guidance for both the manufacturers and competent authorities to manage cases efficiently and appropriately. India has emerged as one of the leaders in pharmaceutical industry. Like many other amendments in Drugs and Cosmetics Act that ...

  15. Electron Cyclotron Current Drive Compensation of the Bootstrap Current in Quasi-symmetric Reactor Devices

    International Nuclear Information System (INIS)

    Margalet, S. D.; Cooper, W. A.; Volpe, F.; Castejon, F.

    2005-01-01

    In magnetic confinement devices, the inhomogeneity of the confining magnetic field along a magnetic field line generates the trapping of particles within local magnetic wells. One of the consequences of the trapped particles is the generation of a current, known as the bootstrap current (BC), whose direction depends on the nature of the magnetic trapping. The BC provides an extra contribution to the poloidal component of the confining magnetic field. The variation of the poloidal component produces the alteration of the winding of the magnetic field lines around the flux surfaces quantified by the rotational transform. When reaches low rational values, it can trigger the generation of ideal MHD instabilities. Therefore, the BC may be responsible for the destabilisation of the configuration [1]. Having established the potentially dangerous implication of the BC, principally, in reactor prototypes, a method to compensate its harmful effects is proposed. It consists of the modelling of the current driven by externally launched ECWs within the plasma to compensate the effects of the BC. This method is flexible enough to allow the identification of the appropriate scenarios in which to generate the required CD depending on the nature of the confining magnetic field and the specific plasma parameters of the configuration. Both the BC and the CD calculations are included in a self-consistent scheme which leads to the computation of a stable BC+CD-consistent MHD equilibrium. This procedure is applied in this paper to simulate the required CD to stabilise a QAS and a QHS reactor prototypes. The estimation of the input power required and the effect of the driven current on the final equilibrium of the system is performed for several relevant scenarios and wave polarisations providing various options of stabilising driven currents. (Author)

  16. Development of a wireless blood pressure measuring device with smart mobile device.

    Science.gov (United States)

    İlhan, İlhan; Yıldız, İbrahim; Kayrak, Mehmet

    2016-03-01

    Today, smart mobile devices (telephones and tablets) are very commonly used due to their powerful hardware and useful features. According to an eMarketer report, in 2014 there were 1.76 billion smartphone users (excluding users of tablets) in the world; it is predicted that this number will rise by 15.9% to 2.04 billion in 2015. It is thought that these devices can be used successfully in biomedical applications. A wireless blood pressure measuring device used together with a smart mobile device was developed in this study. By means of an interface developed for smart mobile devices with Android and iOS operating systems, a smart mobile device was used both as an indicator and as a control device. The cuff communicating with this device through Bluetooth was designed to measure blood pressure via the arm. A digital filter was used on the cuff instead of the traditional analog signal processing and filtering circuit. The newly developed blood pressure measuring device was tested on 18 patients and 20 healthy individuals of different ages under a physician's supervision. When the test results were compared with the measurements made using a sphygmomanometer, it was shown that an average 93.52% accuracy in sick individuals and 94.53% accuracy in healthy individuals could be achieved with the new device. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  17. Measurement of MOS current mismatch in the weak inversion region

    International Nuclear Information System (INIS)

    Forti, F.; Wright, M.E.

    1994-01-01

    The MOS transistor matching properties in the weak inversion region have not received, in the past, the attention that the mismatch in the strong inversion region has. The importance of weak inversion biased transistors in low power CMOS analog systems calls for more extensive data on the mismatch in this region of operation. The study presented in this paper was motivated by the need of controlling the threshold matching in a low power, low noise amplifier discriminator circuit used in a silicon radiation detector read-out, where both the transistor dimensions and the currents had to be kept to a minimum. The authors have measured the current matching properties of MOS transistors operated in the weak inversion region. They measured a total of about 1,400 PMOS and NMOS transistors produced in four different processes and report here the results in terms of mismatch dependence on current density, device dimensions, and substrate voltage, without using any specific model for the transistor

  18. Device to measure elastic modulus of superconducting windings

    CERN Multimedia

    CERN PhotoLab

    1979-01-01

    This device was made to measure elastic modulus of the Po dipole superconducting coils. More elaborated devices, but based on the same concept, were later used to measure the apparent elastic moduli of the LHC superconducting magnet coils. See also 7903547X, 7901386.

  19. Device to measure level in a steam drum of NPP

    International Nuclear Information System (INIS)

    Vinogradov, Yu.A.

    1988-01-01

    Gravitation-hydrostatic device for measuring coolant level in a steam drum of NPP is described. The device enables to improve the accuracy and sensitivity of measuring coolant level above and below the submerged perforated sheet of the steam drum and decrease the amount of levelling vessels in the unit by 50%. 1 fig

  20. Logarithmic current-measuring transistor circuits

    DEFF Research Database (Denmark)

    Højberg, Kristian Søe

    1967-01-01

    Describes two transistorized circuits for the logarithmic measurement of small currents suitable for nuclear reactor instrumentation. The logarithmic element is applied in the feedback path of an amplifier, and only one dual transistor is used as logarithmic diode and temperature compensating...... transistor. A simple one-amplifier circuit is compared with a two-amplifier system. The circuits presented have been developed in connexion with an amplifier using a dual m.o.s. transistor input stage with diode-protected gates....

  1. Ionic current devices-Recent progress in the merging of electronic, microfluidic, and biomimetic structures.

    Science.gov (United States)

    Koo, Hyung-Jun; Velev, Orlin D

    2013-05-09

    We review the recent progress in the emerging area of devices and circuits operating on the basis of ionic currents. These devices operate at the intersection of electrochemistry, electronics, and microfluidics, and their potential applications are inspired by essential biological processes such as neural transmission. Ionic current rectification has been demonstrated in diode-like devices containing electrolyte solutions, hydrogel, or hydrated nanofilms. More complex functions have been realized in ionic current based transistors, solar cells, and switching memory devices. Microfluidic channels and networks-an intrinsic component of the ionic devices-could play the role of wires and circuits in conventional electronics.

  2. Development and optimization of a device for diferencial pressure measurement

    International Nuclear Information System (INIS)

    Santarine, G.A.

    1980-01-01

    The measurements of reduced values of diferencial pressure, are studied. Several situations are described where the diferencial pressure accurate measurement is necessary in routine works in the Thermohydraulic Laboratory, as well as, the major pressure measurement devices and their respective range are studied. The development of a device for diferencial pressure measurement followed by the design development of the calibration bench covering the foreseen range, start up tests realization, optimization, calibration, performance analysis and conclusions, is showed. (Author) [pt

  3. Correlation Measurements on Small Mobile Devices

    DEFF Research Database (Denmark)

    Yanakiev, Boyan Radkov; Nielsen, Jesper Ødum; Christensen, Morten

    2012-01-01

    Here, analysis of the antenna correlation at the design stage is done, with focus on measurement techniques. Various theoretical definitions of correlations are used with the corresponding measured data required. The problems related to the coaxial measurement cables, when calculating correlation...

  4. Educational Behavior Apps and Wearable Devices: Current Research and Prospects

    Science.gov (United States)

    Lowe, Heather

    2016-01-01

    Dartmouth and MIT have developed educational behavior apps and wearable devices that collect contiguous streams of data from student users. Given the consent of the user, the app collects information about a student's physical activity, sleep patterns, and location to form conjectures about social and academic behavior. These apps have the…

  5. First principle leakage current reduction technique for CMOS devices

    CSIR Research Space (South Africa)

    Tsague, HD

    2015-12-01

    Full Text Available This paper presents a comprehensive study of leakage reduction techniques applicable to CMOS based devices. In the process, mathematical equations that model the power-performance trade-offs in CMOS logic circuits are presented. From those equations...

  6. High current density M-type cathodes for vacuum electron devices

    International Nuclear Information System (INIS)

    Li Ji; Yu Zhiqiang; Shao Wensheng; Zhang Ke; Gao Yujuan; Yuan Haiqing; Wang Hui; Huang Kaizhi; Chen Qilue; Yan Suqiu; Cai Shaolun

    2005-01-01

    We investigated high current density emission capabilities of M-type cathodes used for vacuum electron devices (VEDs). The experimental results of emission and lifetime evaluating in both close-spaced diode structure and electron gun testing vehicles are given. Emission current densities measured in the diode structure at 1020 deg. C Br in the CW mode were above 10 A/cm 2 ; while in electron gun testing vehicles, emission current densities were above 8 A/cm 2 in CW mode and above 32 A/cm 2 in pulsed mode, respectively. The current density above 94 A/cm 2 has been acquired in no. 0306 electron gun vehicle while the practical temperature is 1060 deg. C Br . For a comparison some of the data from I-scandate cathodes are presented. Finally, several application examples in practical travelling wave tubes (TWTs) and multi beam klystrons (MBKs) are also reported

  7. Optical vibration measurement of mechatronics devices

    Science.gov (United States)

    Yanabe, Shigeo

    1993-09-01

    An optical vibration measuring system which enables to detect both linear and angular displacement of 25 nm and 5 prad was developed. The system is mainly composed of a He-Ne laser, a displacement detecting photo-diode and lenses, and has linear and angular displacement magnification mechanism using two different principles of optical lever. The system was applied to measure vibrational characteristics of magnetic head slider of hard disk drives and to measure stator teeth driving velocities of ultrasonic motor.

  8. Test device for measuring permeability of a barrier material

    Science.gov (United States)

    Reese, Matthew; Dameron, Arrelaine; Kempe, Michael

    2014-03-04

    A test device for measuring permeability of a barrier material. An exemplary device comprises a test card having a thin-film conductor-pattern formed thereon and an edge seal which seals the test card to the barrier material. Another exemplary embodiment is an electrical calcium test device comprising: a test card an impermeable spacer, an edge seal which seals the test card to the spacer and an edge seal which seals the spacer to the barrier material.

  9. Methods of radon measurement and devices

    International Nuclear Information System (INIS)

    Miles, J.

    2004-01-01

    The following topics and instrumentation are discussed: The quantity to be measured; Active measurement methods (scintillation cells, ionisation chambers, electrostatic collection of decay products); Passive measurement methods (charcoal detectors; electret ion chambers; etched track detectors); and Detector considerations for large-scale surveys ('always on' or 'switchable' detectors?; response to radon-220; avoidance of electrostatic effects; quality assurance for passive radon detectors; quality control within the laboratory; external quality assurance; detectors need to be easily deliverable). It is concluded that the ideal detector for large scale surveys of radon in houses is a small, closed detector in a conducting holder which excludes radon-220, supported by rigorous quality assurance procedures. (P.A.)

  10. Fluxgate Magnetic Sensor and Its Application for Current Measurement

    International Nuclear Information System (INIS)

    Mitra-Djamal

    2007-01-01

    Conventionally electric current can be measured by connecting the instrument serially on the circuit. This method has disadvantage because its disturb the measured current flow. By using a magnetic sensor, current can be measured without disturbing the current flow, because it just measures the magnetic field of the measured current. This paper shows the use of fluxgate magnetic sensor for current measurement. It is shown that the sensor can measure widely range of current with resolution ≤ 2 %. (author)

  11. Method and apparatus to trigger superconductors in current limiting devices

    Science.gov (United States)

    Yuan, Xing; Hazelton, Drew Willard; Walker, Michael Stephen

    2004-10-26

    A method and apparatus for magnetically triggering a superconductor in a superconducting fault current limiter to transition from a superconducting state to a resistive state. The triggering is achieved by employing current-carrying trigger coil or foil on either or both the inner diameter and outer diameter of a superconductor. The current-carrying coil or foil generates a magnetic field with sufficient strength and the superconductor is disposed within essentially uniform magnetic field region. For superconductor in a tubular-configured form, an additional magnetic field can be generated by placing current-carrying wire or foil inside the tube and along the center axial line.

  12. Radiation protection measuring device SSM-1

    International Nuclear Information System (INIS)

    Anon.

    1988-01-01

    Product information from the producer on a universal measuring instrument for alpha, beta and gamma radiation designed for stationary and field use by military, police and fire brigades. 4 figs. (qui)

  13. Methods and Devices used to Measure Friction

    DEFF Research Database (Denmark)

    Jeswiet, Jack; Arentoft, Mogens; Henningsen, Poul

    2004-01-01

    . To gain a good understanding of the mechanisms at the interface and to be able to verify the friction and tribology models that exist, friction sensors are needed. Designing sensors to measure friction-stress in metal working has been pursued by many researchers. This paper surveys methods, which have...... been tried in the past and discusses some of the recent sensor designs, which can now be used to measure Friction in both production situations and for research purposes....

  14. Memory-assisted measurement-device-independent quantum key distribution

    Science.gov (United States)

    Panayi, Christiana; Razavi, Mohsen; Ma, Xiongfeng; Lütkenhaus, Norbert

    2014-04-01

    A protocol with the potential of beating the existing distance records for conventional quantum key distribution (QKD) systems is proposed. It borrows ideas from quantum repeaters by using memories in the middle of the link, and that of measurement-device-independent QKD, which only requires optical source equipment at the user's end. For certain memories with short access times, our scheme allows a higher repetition rate than that of quantum repeaters with single-mode memories, thereby requiring lower coherence times. By accounting for various sources of nonideality, such as memory decoherence, dark counts, misalignment errors, and background noise, as well as timing issues with memories, we develop a mathematical framework within which we can compare QKD systems with and without memories. In particular, we show that with the state-of-the-art technology for quantum memories, it is potentially possible to devise memory-assisted QKD systems that, at certain distances of practical interest, outperform current QKD implementations.

  15. Plasma position from ring current measurements in Extrap T1

    International Nuclear Information System (INIS)

    Brunsell, P.; Jin Li.

    1989-11-01

    The inductive coupling between the plasma and the four octupole field coils in the Extrap T1 device is utilized as a means of estimating the plasma position. The current in each octupole ring as well as the plasma current is measured by a Rogowski coil and the ring - plasma mutual inductance is then computed assuming axisymmetric plasma displacements. The obtained position is in agreement with internal magnetic probe measurements. The time - evolution of the plasma position for different external vertical and toroidal field strengths is studied. For the present discharge parameter a vertical field of about .008 T is found to give an almost radially stationary plasma. The results are compared with a simple equilibrium model

  16. Small Device For Short-Range Antenna Measurements Using Optics

    DEFF Research Database (Denmark)

    Yanakiev, Boyan Radkov; Nielsen, Jesper Ødum; Christensen, Morten

    2011-01-01

    This paper gives a practical solution for implementing an antenna radiation pattern measurement device using optical fibers. It is suitable for anechoic chambers as well as short range channel sounding. The device is optimized for small size and provides a cheap and easy way to make optical antenna...

  17. Miniature ingestible telemeter devices to measure deep-body temperature

    Science.gov (United States)

    Pope, J. M.; Fryer, T. B. (Inventor)

    1976-01-01

    A telemetry device comprised of a pill-size ingestible transmitter developed to obtain deep body temperature measurements of a human is described. The device has particular utility in the medical field where deep body temperatures provide an indication of general health.

  18. Video integrated measurement system. [Diagnostic display devices

    Energy Technology Data Exchange (ETDEWEB)

    Spector, B.; Eilbert, L.; Finando, S.; Fukuda, F.

    1982-06-01

    A Video Integrated Measurement (VIM) System is described which incorporates the use of various noninvasive diagnostic procedures (moire contourography, electromyography, posturometry, infrared thermography, etc.), used individually or in combination, for the evaluation of neuromusculoskeletal and other disorders and their management with biofeedback and other therapeutic procedures. The system provides for measuring individual diagnostic and therapeutic modes, or multiple modes by split screen superimposition, of real time (actual) images of the patient and idealized (ideal-normal) models on a video monitor, along with analog and digital data, graphics, color, and other transduced symbolic information. It is concluded that this system provides an innovative and efficient method by which the therapist and patient can interact in biofeedback training/learning processes and holds considerable promise for more effective measurement and treatment of a wide variety of physical and behavioral disorders.

  19. Dose-current discharge correlation analysis in a Mather type Plasma Focus device for medical applications

    Science.gov (United States)

    Sumini, M.; Mostacci, D.; Tartari, A.; Mazza, A.; Cucchi, G.; Isolan, L.; Buontempo, F.; Zironi, I.; Castellani, G.

    2017-11-01

    In a Plasma Focus device the plasma collapses into the pinch where it reaches thermonuclear conditions for a few tens of nanoseconds, becoming a multi-radiation source. The nature of the radiation generated depends on the gas filling the chamber and the device working parameters. The self-collimated electron beam generated in the backward direction with respect to the plasma motion is one of the main radiation sources of interest also for medical applications. The electron beam may be guided against a high Z material target to produce an X-ray beam. This technique offers an ultra-high dose rate source of X-rays, able to deliver during the pinch a massive dose (up to 1 Gy per discharge for the PFMA-3 test device), as measured with EBT3 GafchromicⒸfilm tissue equivalent dosimeters. Given the stochastic behavior of the discharge process, a reliable on-line estimate of the dose-delivered is a very challenging task, in some way preventing a systematic application as a potentially interesting therapy device. This work presents an approach to linking the dose registered by the EBT3 GafchromicⒸfilms with the information contained in the signal recorded during the current discharge process. Processing the signal with the Wigner-Ville distribution, a spectrogram was obtained, displaying the information on intensity at various frequency scales, identifying the band of frequencies representative of the pinch events and define some patterns correlated with the dose.

  20. Liquid temperature measuring method and device therefor

    Energy Technology Data Exchange (ETDEWEB)

    Maruyama, Fumi; Karasawa, Hirokazu

    1995-06-02

    In the present invention, temperature of liquid metal in coolants in an FBR type reactor can accurately be measured at rapid response time. Namely, ultrasonic waves are emitted from an ultrasonic wave sensor disposed in the air to a guide wave tube. Ultrasonic waves are reflected at reflection plates disposed at front and back or upper and lower portions of a small hole disposed to the wave guide tube. The reflected waves are received by the sensor described above. The difference of the reaching time of the reflected waves from the reflecting plates disposed at the front and the back or the upper and lower portions is measured. The speed of sounds in this case is determined based on the size of the small hole and the distance of the upper and the lower reflection plates. The speed of sounds is determined by the formula below: V(m/s) = 2500 - 0.52 T, where T: temperature. The temperature of the liquid can easily be calculated based on the formula. Accordingly, since the speed of the ultrasonic waves from their emission to the reception is msec order, and the processing of the signals are simple, the temperature can be measured at a response time of several msecs. In addition, since the ultrasonic wave sensor is disposed at the outside of the reactor, no special countermeasure for environmental circumstances is necessary, to improve maintenance ability. (I.S.).

  1. Characteristics of the Current-Controlled Phase Transition of VO2 Microwires for Hybrid Optoelectronic Devices

    Directory of Open Access Journals (Sweden)

    Arash Joushaghani

    2015-08-01

    Full Text Available The optical and electrical characteristics of the insulator-metal phase transition of vanadium dioxide (VO2 enable the realization of power-efficient, miniaturized hybrid optoelectronic devices. This work studies the current-controlled, two-step insulator-metal phase transition of VO2 in varying microwire geometries. Geometry-dependent scaling trends extracted from current-voltage measurements show that the first step induced by carrier injection is delocalized over the microwire, while the second, thermally-induced step is localized to a filament about 1 to 2 μm wide for 100 nm-thick sputtered VO2 films on SiO2. These effects are confirmed by direct infrared imaging, which also measures the change in optical absorption in the two steps. The difference between the threshold currents of the two steps increases as the microwires are narrowed. Micron- and sub-micron-wide VO2 structures can be used to separate the two phase transition steps in photonic and electronic devices.

  2. Modified sine bar device measures small angles with high accuracy

    Science.gov (United States)

    Thekaekara, M.

    1968-01-01

    Modified sine bar device measures small angles with enough accuracy to calibrate precision optical autocollimators. The sine bar is a massive bar of steel supported by two cylindrical rods at one end and one at the other.

  3. Zero G Mass Measurement Device (ZGMMD), Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — The Zero G Mass Measurement Device (ZGMMD) will provide the ability to quantify the mass of objects up to 2,000 grams, including live animal specimens in a zero G...

  4. Gate current for p+-poly PMOS devices under gate injection conditions

    NARCIS (Netherlands)

    Hof, A.J.; Holleman, J.; Woerlee, P.H.

    2001-01-01

    In current CMOS processing both n+-poly and p+-poly gates are used. The I-V –relationship and reliability of n+-poly devices are widely studied and well understood. Gate currents and reliability for p+-poly PMOS devices under gate injection conditions are not well understood. In this paper, the

  5. Evaluating the Consistency of Current Mainstream Wearable Devices in Health Monitoring: A Comparison Under Free-Living Conditions.

    Science.gov (United States)

    Wen, Dong; Zhang, Xingting; Liu, Xingyu; Lei, Jianbo

    2017-03-07

    ). Although wearable devices are developing rapidly, the current mainstream devices are only reliable in measuring the number of steps and distance, which can be used as health assessment indicators. However, the measurement consistencies of activity duration, EE, sleep quality, and so on, are still inadequate, which require further investigation and improved algorithms. ©Dong Wen, Xingting Zhang, Xingyu Liu, Jianbo Lei. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 07.03.2017.

  6. Magnitude of the current in 2D interlayer tunneling devices.

    Science.gov (United States)

    Feenstra, Randall M; de la Barrera, Sergio C; Li, Jun; Nie, Yifan; Cho, Kyeongjae

    2018-01-15

    Using the Bardeen tunneling method with first-principles wave functions, computations are made of the tunneling current in graphene/hexagonal-boron-nitride/graphene (G/h-BN/G) vertical structures. Detailed comparison with prior experimental results is made, focusing on the magnitude of the achievable tunnel current. With inclusion of the effects of translational and rotational misalignment of the graphene and the h-BN, predicted currents are found to be about 15×  larger than experimental values. A reduction in this discrepancy, to a factor of 2.5×, is achieved by utilizing a realistic size for the band gap of the h-BN, hence affecting the exponential decay constant for the tunneling.

  7. Accuracy of portable devices in measuring peak cough flow

    International Nuclear Information System (INIS)

    Kulnik, Stefan Tino; Kalra, Lalit; MacBean, Victoria; Birring, Surinder Singh; Moxham, John; Rafferty, Gerrard Francis

    2015-01-01

    Peak cough flow (PCF) measurements can be used as indicators of cough effectiveness. Portable peak flow meters and spirometers have been used to measure PCF, but little is known about their accuracy compared to pneumotachograph systems. The aim of this study was to compare the accuracy of four portable devices (Mini–Wright and Assess peak flow meters, SpiroUSB and Microlab spirometers) in measuring PCF with a calibrated laboratory based pneumotachograph system. Twenty healthy volunteers (mean (SD) age 45 (16) years) coughed through a pneumotachograph connected in series with each portable device in turn, and the differences in PCF readings were analysed. In addition, mechanically generated flow waves of constant peak flow were delivered through each device both independently and when connected in series with the pneumotachograph. Agreement between PCF readings obtained with the pneumotachograph and the portable devices was poor. Peak flow readings were on average lower by approximately 50 L min −1 when measured using the portable devices; 95% limits of agreement spanned approximately 150 L min −1 . The findings highlight the potential for inaccuracy when using portable devices for the measurement of PCF. Depending on the measurement instrument used, absolute values of PCF reported in the literature may not be directly comparable. (paper)

  8. Device-Independent Certification of a Nonprojective Qubit Measurement

    Science.gov (United States)

    Gómez, Esteban S.; Gómez, Santiago; González, Pablo; Cañas, Gustavo; Barra, Johanna F.; Delgado, Aldo; Xavier, Guilherme B.; Cabello, Adán; Kleinmann, Matthias; Vértesi, Tamás; Lima, Gustavo

    2016-12-01

    Quantum measurements on a two-level system can have more than two independent outcomes, and in this case, the measurement cannot be projective. Measurements of this general type are essential to an operational approach to quantum theory, but so far, the nonprojective character of a measurement can only be verified experimentally by already assuming a specific quantum model of parts of the experimental setup. Here, we overcome this restriction by using a device-independent approach. In an experiment on pairs of polarization-entangled photonic qubits we violate by more than 8 standard deviations a Bell-like correlation inequality that is valid for all sets of two-outcome measurements in any dimension. We combine this with a device-independent verification that the system is best described by two qubits, which therefore constitutes the first device-independent certification of a nonprojective quantum measurement.

  9. Radon Measurement Proficiency (RMP) Program methods and devices

    International Nuclear Information System (INIS)

    Harrison, J.; Hoornbeek, J.; Jalbert, P.; Sensintaffar, E.; Hopper, R.

    1991-01-01

    The US EPA developed the voluntary Radon Measurement Proficiency Program in 1986 in response to a Federal and State need for measurement services firms to demonstrate their proficiency with radon measurement methods and devices. Since that time, the program has set basic standards for the radon measurement industry. The program has grown dramatically since its inception. In 1986, fewer than 50 companies participated in the program. By 1989, more than 5,000 companies were participating. Participants represent firms with an analytical capability as well as firms that rely upon another firm for analysis service. Since the beginning of the RMP Program, the Agency has learned a great deal about radon measurement methods and devices. This paper reviews the measurement devices used in the program and what the EPA has learned about them since the program's inception. Performance data from the RMP Program are used to highlight relevant findings

  10. 40 CFR 1065.275 - N2O measurement devices.

    Science.gov (United States)

    2010-07-01

    ... measurement devices. (a) General component requirements. We recommend that you use an analyzer that meets the... functions of other gaseous measurements and the engine's known or assumed fuel properties. The target value... gaseous measurements. The target value for any compensation algorithm is 0.0% (that is, no bias high and...

  11. OVERVIEW OF NEUTRON MEASUREMENTS IN JET FUSION DEVICE.

    Science.gov (United States)

    Batistoni, P; Villari, R; Obryk, B; Packer, L W; Stamatelatos, I E; Popovichev, S; Colangeli, A; Colling, B; Fonnesu, N; Loreti, S; Klix, A; Klosowski, M; Malik, K; Naish, J; Pillon, M; Vasilopoulou, T; De Felice, P; Pimpinella, M; Quintieri, L

    2017-10-05

    The design and operation of ITER experimental fusion reactor requires the development of neutron measurement techniques and numerical tools to derive the fusion power and the radiation field in the device and in the surrounding areas. Nuclear analyses provide essential input to the conceptual design, optimisation, engineering and safety case in ITER and power plant studies. The required radiation transport calculations are extremely challenging because of the large physical extent of the reactor plant, the complexity of the geometry, and the combination of deep penetration and streaming paths. This article reports the experimental activities which are carried-out at JET to validate the neutronics measurements methods and numerical tools used in ITER and power plant design. A new deuterium-tritium campaign is proposed in 2019 at JET: the unique 14 MeV neutron yields produced will be exploited as much as possible to validate measurement techniques, codes, procedures and data currently used in ITER design thus reducing the related uncertainties and the associated risks in the machine operation. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  12. Measuring device for soft X-rays

    International Nuclear Information System (INIS)

    Dissing, E.

    1978-09-01

    An instrument for the measurement of the absorbed energy per unit area of diagnostic X-rays in soft human tissue was developed. The instrument is intended for dosimetry applications in the field of dental and small skeleton radiography and for mammography. The detector assembly consists of a Polyvinyltoluene scintillator 2.54 diametre x 5.08 cm CsSb semitransparent head-on vacuum phototube. Polyvinyltoluene being a pure hydrocarbon may be considered a good representative material of human soft tissue concerning the absorption of X-rays. In the photon energy range of interest, 5 - 40 keV, the mass energy absorption coefficient for muscle tissue and for PVT differ about a factor 2 due to the considerable content of Oxygen in muscle tissue. This is to some extend reflected in the photon energy response characteristic for the instrument. For human adipose, the characteristic is practically flat from 5- 40 keV. The instrument is integrating the absorbed power per unit area and the digital display shows Joules/m 2 . The range for the instrument is from 000.1 μJ/m 2 to 19.99 J/m 2 (absorbed energy in 5 cm tissue). (author)

  13. Device for measuring a burnup degree

    International Nuclear Information System (INIS)

    Ito, Toshiaki; Goto, Seiichiro

    1979-01-01

    Purpose: To measure the burnup degree at high efficiency and accuracy. Constitution: The outer metal wall of fuel assemblies is heated under gamma radiation with long half life gamma rays in inverse proportion to the burnup degree and issues infrared radiation in proportion to the intensity of the gamma rays. An image pick-up tube is opposed to one surface of the fuel assemblies to detect the radiated infrared rays. Since the output signal from the pick-up tube is subjected to the absorptive damping by the distance between the pick-up tube and the fuel assembly, as well as water filled in the gap therebetween, it is corrected through a main amplifier comprising a signal correction circuit composed of a characteristic section inverse to the absorption property and a characteristic section inverse to the square of the distance. The corrected output signal is displayed on a display unit such as CRT or recorded in a film or a magnetic tape. (Furukawa, Y.)

  14. Device for the alternative option of temperature measurement

    Science.gov (United States)

    Jargus, Jan; Nedoma, Jan; Fajkus, Marcel; Novak, Martin; Cubik, Jakub; Cvejn, Daniel; Vasinek, Vladimir

    2017-10-01

    Polydimethylsiloxane (PDMS) has good optical properties, and its composition offers the possibility of use in many applications (industry, security device, medicine applications and etc.). We focused on the alternative option of temperature measurement in this article. Our approach is based on measuring changes of chromaticity correlated temperature corresponding to changes in temperature. Described device uses an optical fiber with a defined layer of PDMS and luminophore and we assume that it can find use also in the field of security. The article describes the process of making the prototype of the device and its verification based on laboratory results. The measured temperature depends mainly on the type of optical fiber and the measured temperature range is determined by the thermal resistance of used optical fiber. Using a calibration measurement can determine the value of temperature with an accuracy of +/- 2,5 %.

  15. Current sheath curvature correlation with the neon soft x-ray emission from plasma focus device

    International Nuclear Information System (INIS)

    Zhang, T; Lin, X; Chandra, K A; Tan, T L; Springham, S V; Patran, A; Lee, P; Lee, S; Rawat, R S

    2005-01-01

    The insulator sleeve length is one of the major parameters that can severely affect the neon soft x-ray yield from a plasma focus. The effect of the insulation sleeve length on various characteristic timings of plasma focus discharges and hence the soft x-ray emission characteristics has been investigated using a resistive divider. The pinhole images and laser shadowgraphy are used to explain the observed variation in the average soft x-ray yield (measured using a diode x-ray spectrometer) with variation of the insulator sleeve length. We have found that for a neon filled plasma focus device the change in insulator sleeve length changes the current sheath curvature angle and thus the length of the focused plasma column. The optimized current sheath curvature angle is found to be between 39 0 and 41 0 , at the specific axial position of 6.2-9.3 cm from the cathode support plate, for our 3.3 kJ plasma focus device. A strong dependence of the neon soft x-ray yield on the current sheath curvature angle has thus been reported

  16. Impacts of Ripple Current to the Loading and Lifetime of Power Semiconductor Device

    DEFF Research Database (Denmark)

    Ma, Ke; Choi, Uimin; Blaabjerg, Frede

    2017-01-01

    The thermal loading of power electronics devices is determined by many factors and has being a crucial design consideration because it is closely related to the reliability and cost of the converter system. In this paper the impacts of the ripple current to the loss and thermal loading, as well...... as reliability performances of power devices are comprehensively investigated and tested. It is concluded that the amplitude of ripple current may modify the loss and thermal loading of the power devices, especially under the conditions of converter with low power output, and thus the lifetime of devices could...

  17. Device independent quantum key distribution secure against coherent attacks with memoryless measurement devices

    International Nuclear Information System (INIS)

    McKague, Matthew

    2009-01-01

    Device independent quantum key distribution (QKD) aims to provide a higher degree of security than traditional QKD schemes by reducing the number of assumptions that need to be made about the physical devices used. The previous proof of security by Pironio et al (2009 New J. Phys. 11 045021) applies only to collective attacks where the state is identical and independent and the measurement devices operate identically for each trial in the protocol. We extend this result to a more general class of attacks where the state is arbitrary and the measurement devices have no memory. We accomplish this by a reduction of arbitrary adversary strategies to qubit strategies and a proof of security for qubit strategies based on the previous proof by Pironio et al and techniques adapted from Renner.

  18. Development of high impedance measurement system for water leakage detection in implantable neuroprosthetic devices.

    Science.gov (United States)

    Yousif, Aziz; Kelly, Shawn K

    2016-08-01

    There has been a push for a greater number of channels in implantable neuroprosthetic devices; but, that number has largely been limited by current hermetic packaging technology. Microfabricated packaging is becoming reality, but a standard testing system is needed to prepare these devices for clinical trials. Impedance measurements of electrodes built into the packaging layers may give an early warning of device failure and predict device lifetime. Because the impedance magnitudes of such devices can be on the order of gigaohms, a versatile system was designed to accommodate ultra-high impedances and allow future integrated circuit implementation in current neural prosthetic technologies. Here we present the circuitry, control software, and preliminary testing results of our designed system.

  19. Backscattering measurement device for measuring the thickness of a layer

    International Nuclear Information System (INIS)

    Weinstock, J.; Lieber, D.; Hay, W.D.

    1978-01-01

    There is provided for a measuring wheel on the run of which backscattering probes are mounted, serving for irradiation and measurement of the radiation reflected from a strip of substrate tape coated e.g. with Au. The probes are of the model HH-3 of Unit Process Assemblies Inc. The material strip is guided on the outside of the wheel run. The measuring wheel is rotating with such speed that the tangential velocity of a point on the run is equal to the speed of the strip. Therefore the movement of the strip need not be stopped during measurement (on-line measurement). (DG) [de

  20. Nuclear-burst strength detecting and measuring device

    International Nuclear Information System (INIS)

    Balut, J.A.L.G.; Lemaire, P.E.G.K.; Loisy, C.M.

    1976-01-01

    A continuous-operation automatic device is described for detection and accurate measurement of the strength of a burst generating an emission from luminous or infrared sources. This device characterizes and analyzes the maxima and minima of a ''thermal flux/time'' curve. The device comprises a master time element and an assembly of photoelectric detectors, an electronic processing system coupled to the detectors, and a mechanical system securing the rigidity and positioning of the photoelectric detector assembly with respect to an octahedral prism based on a horizontal plane

  1. An Innovative Flow-Measuring Device: Thermocouple Boundary Layer Rake

    Science.gov (United States)

    Hwang, Danny P.; Fralick, Gustave C.; Martin, Lisa C.; Wrbanek, John D.; Blaha, Charles A.

    2001-01-01

    An innovative flow-measuring device, a thermocouple boundary layer rake, was developed. The sensor detects the flow by using a thin-film thermocouple (TC) array to measure the temperature difference across a heater strip. The heater and TC arrays are microfabricated on a constant-thickness quartz strut with low heat conductivity. The device can measure the velocity profile well into the boundary layer, about 65 gm from the surface, which is almost four times closer to the surface than has been possible with the previously used total pressure tube.

  2. Propagation Measurements for Device-to-Device Communication in Forest Terrain

    DEFF Research Database (Denmark)

    Hejselbæk, Johannes; Nielsen, Jesper Ødum; Drewes, Christian

    2018-01-01

    In this paper, we present a measurement campaign conducted in forest terrain with focus on path-loss. The aim of the measurement campaign is to study the coverage in a Device-to-Device (D2D) communication scenario. The measurement campaign was conducted in the LTE band 8 at 917.5 MHz...... with measurement ranges extending to more than 2.5 km. The measurements have been conducted using a purpose-developed measurement system with a dynamic range of 180 dB. The measurements showed that a D2D system with transmit and receive antenna at heights of 1.5 m could achieve a range of approximately 2 km using...

  3. Device for measuring the tritium concentration in a measuring gas

    International Nuclear Information System (INIS)

    Koran, P.

    1987-01-01

    The measuring gas is brought into contact via a measuring gas path with a diaphragm permeable to water, which separates the measuring gas path from a counter gas path leading to a proportional detector. The measuring gas path and the counter gas path are in counterflow in the area of diaphragm. The preferably hose diaphragm consists of a well-known ion exchange material, which can be used for gas drying purposes, which is permeable to water and tritium compounds similar to water, but is impermeable to other gases and liquids contained in air, particularly rare gases. In this way, the tritium concentration can be measured with great rare gas suppression. (orig./HP) [de

  4. Study of efficiency and current use of respiratory protective devices

    International Nuclear Information System (INIS)

    Linauskas, S.H.; Kalos, F.

    1984-12-01

    The efficiency of a helmet-type powered air-purifying respirator was studied in the work environment of an underground uranium mine. Workplace protection factors (WPF) were measured, using approximately 30-minute air sampling periods and repeated sampling throughout the work shifts for six miners during three shifts each. The WPF of the modified Racal AH-5 helmet was 17.8 (geometric mean) with new filters in the respirator; the WPF dropped to 5.3 when four-day-old filters were used. The WPF results were lognormally distributed. The test population appeared to be a homogeneous set: there was no significant difference in the mean WPFs of individual test subjects. The study could not resolve any effect of elapsed time within the work shift or of ambient Working Level on the WPF. The actual use levels and the problems associated with respirator use were studied at the same time, principally via a user survey. The respirators have a high level of acceptance among the user population of underground uranium miners and 72.5 percent of the 207 survey respondents stated six or more hours use with the visor in the correct 'down' position per shift

  5. Measurement Devices and the Psychophysiology of Consumer Behaviour

    DEFF Research Database (Denmark)

    Schwarzkopf, Stefan

    2015-01-01

    of the type of subjectivity that underlies consumer behaviour. I argue instead that a posthuman view of the relationship between brain, mind and behaviour underpinned neurophysiological research into consumers from its very beginning in the late nineteenth century. By tracing the biopolitical potentialities...... of neuromarketing back to the Fin-de-Siècle neurophysiological laboratory, I show that consumers' bodies and later on their brains became reconfigured as part of a dispositif made up of laboratory-based artefacts (measurement devices) and new ways of seeing the human brain and human behaviour. This dispositif......From the 1890s, psychophysiological measurement devices have played an important, but as yet under-theorized role in marketing and consumer research. Because of the recent advances made in neuromarketing, it is often assumed that these measurement devices ushered in a radically new understanding...

  6. SHORT COMMUNICATION: Time measurement device with four femtosecond stability

    Science.gov (United States)

    Panek, Petr; Prochazka, Ivan; Kodet, Jan

    2010-10-01

    We present the experimental results of extremely precise timing in the sense of time-of-arrival measurements in a local time scale. The timing device designed and constructed in our laboratory is based on a new concept using a surface acoustic wave filter as a time interpolator. Construction of the device is briefly described. The experiments described were focused on evaluating the timing precision and stability. Low-jitter test pulses with a repetition frequency of 763 Hz were generated synchronously to the local time base and their times of arrival were measured. The resulting precision of a single measurement was typically 900 fs RMS, and a timing stability TDEV of 4 fs was achieved for time intervals in the range from 300 s to 2 h. To our knowledge this is the best value reported to date for the stability of a timing device. The experimental results are discussed and possible improvements are proposed.

  7. Band to Band Tunneling (BBT) Induced Leakage Current Enhancement in Irradiated Fully Depleted SOI Devices

    Science.gov (United States)

    Adell, Phillipe C.; Barnaby, H. J.; Schrimpf, R. D.; Vermeire, B.

    2007-01-01

    We propose a model, validated with simulations, describing how band-to-band tunneling (BBT) affects the leakage current degradation in some irradiated fully-depleted SOI devices. The dependence of drain current on gate voltage, including the apparent transition to a high current regime is explained.

  8. A Novel Electronic Device for Measuring Urine Flow Rate: A Clinical Investigation

    OpenAIRE

    Aliza Goldman; Hagar Azran; Tal Stern; Mor Grinstein; Dafna Wilner

    2017-01-01

    Objective: Currently, most vital signs in the intensive care unit (ICU) are electronically monitored. However, clinical practice for urine output (UO) measurement, an important vital sign, usually requires manual recording of data that is subject to human errors. In this study, we assessed the ability of a novel electronic UO monitoring device to measure real-time hourly UO versus current clinical practice. Design: Patients were connected to the RenalSense Clarity RMS Sensor Kit with a sensor...

  9. Development of a magnetic measurement device for thin ribbon samples

    International Nuclear Information System (INIS)

    Sato, Yuta; Todaka, Takashi; Enokizono, Masato

    2008-01-01

    This paper presents a magnetic measurement device for thin ribbon samples, which are produced by rapid cooling technique. This device enables us to measure magnetic properties easily by only inserting a ribbon sample into a sample holder. The sample holder was made by bakelite to fix any width sample. A long solenoid coil was used to generate a uniform magnetic field and the sample holder was placed at the mid part of the solenoid. The magnetic field strength was measured using a shunt resistor and the magnetic flux density and magnetization in sample ribbons were evaluated by using search coils. The accuracy of measurement was verified with an amorphous metal ribbon sample. Next, we have measured magnetic properties of some magnetic shape memory alloys, which have different compositions. The measured results are compared and we clarified the effect of Sm contents on the magnetic properties

  10. Dynamic pipe control with a multiple digit automatic measuring device

    International Nuclear Information System (INIS)

    Jenzer, P.

    1984-01-01

    With the flow rotating method, thin-walled pipes can be produced with very tight tolerances and high mechanical sturdiness. The measuring device permits a dynamic control of these pipes, the outer diameter of which can lie between 70 and 300 mm, the length between 500 and 2000 mm and the wall thickness between 0,5 and 10 mm. Depending on the pipe type, up to 27 measurements in a maximum of 5 measuring levels are to be controlled. (orig.) [de

  11. Use of piezoelectric multicomponent force measuring devices in fluid mechanics

    Science.gov (United States)

    Richter, A.; Stefan, K.

    1979-01-01

    The characterisitics of piezoelectric multicomponent transducers are discussed, giving attention to the advantages of quartz over other materials. The main advantage of piezoelectric devices in aerodynamic studies is their ability to indicate rapid changes in the values of physical parameters. Problems in the accuracy of measurments by piezoelectric devices can be overcome by suitable design approaches. A practical example is given of how such can be utilized to measure rapid fluctuations of fluid forces exerted on a circular cylinder mounted in a water channel.

  12. The current situation and development of medical device testing institutes in China.

    Science.gov (United States)

    Yang, Xiaofang; Mu, Ruihong; Fan, Yubo; Wang, Chunren; Li, Deyu

    2017-04-01

    This article analyses the current situation and development of Chinese medical device testing institutes from the perspectives of the two most important functions - testing functions and medical device standardization functions. Areas Covered: The objective of the Chinese government regulations for medical device industry is to ensure the safety and effectiveness of medical devices for Chinese patients. To support the regulation system, the Chinese government has established medical device testing institutes at different levels for example, the national, provincial, and municipal levels. These testing institutes also play an important role in technical support during medical device premarket registration and post market surveillance, they are also the vital practitioners of Chinese medical device standardization. Expert Commentary: Chinese medical device testing institutes are technical departments established by government, and serve the regulatory functions of government agency. In recent years, with the rapid development of medical device industry as well as constantly increasing international and domestic medical device market, the importance of medical device testing institute is more prominent, However, there are still some problems unsolved, such as their overall capacity remains to be improved, construction of standardization is to be strengthened, etc.

  13. Memory-assisted measurement-device-independent quantum key distribution

    International Nuclear Information System (INIS)

    Panayi, Christiana; Razavi, Mohsen; Ma, Xiongfeng; Lütkenhaus, Norbert

    2014-01-01

    A protocol with the potential of beating the existing distance records for conventional quantum key distribution (QKD) systems is proposed. It borrows ideas from quantum repeaters by using memories in the middle of the link, and that of measurement-device-independent QKD, which only requires optical source equipment at the user's end. For certain memories with short access times, our scheme allows a higher repetition rate than that of quantum repeaters with single-mode memories, thereby requiring lower coherence times. By accounting for various sources of nonideality, such as memory decoherence, dark counts, misalignment errors, and background noise, as well as timing issues with memories, we develop a mathematical framework within which we can compare QKD systems with and without memories. In particular, we show that with the state-of-the-art technology for quantum memories, it is potentially possible to devise memory-assisted QKD systems that, at certain distances of practical interest, outperform current QKD implementations. (paper)

  14. Comparative analysis of colour change measurement devices in textile industry

    Directory of Open Access Journals (Sweden)

    Paulina Gilewicz

    2014-08-01

    Full Text Available In the paper there is presented a trial of application of new measurement principle of colour change with the use of DigiEye device. Comparison of DigiEye with commonly use in the textile industry spectrophotometer Macbeth 2020 was an aim of determination of relationship between parameters of both measurement systems. Samples for the colour change assessment on both measurement systems were first aged in the Xenotest 150. Ageing process was done according to the method of blues scale. Results were obtained by the colour measurement devices before and after the ageing test each releasing the diaphragms during exposing the examined samples on the light. Result of colour change were obtained in the colour system CIE L*a*b*. The measurements were done for PES fabrics destined on the outer layers of clothing. [b]Keywords[/b]: textiles, spectrophotometer, colorimeter [b][/b

  15. Change of the dominant luminescent mechanism with increasing current density in molecularly doped organic light-emitting devices

    International Nuclear Information System (INIS)

    Zhou Liang; Zhang Hongjie; Meng Qingguo; Liu Fengyi; Yu Jiangbo; Deng Ruiping; Peng Zeping; Li Zhefeng; Guo Zhiyong

    2007-01-01

    We have fabricated and measured a series of electroluminescent devices with the structure of ITO/TPD/Eu(TTA) 3 phen (x):CBP/BCP/ALQ/LiF/Al, where x is the weight percentage of Eu(TTA) 3 phen (from 0% to 6%). At very low current density, carrier trapping is the dominant luminescent mechanism and the 4% doped device shows the highest electroluminescence (EL) efficiency among all these devices. With increasing current density, Foerster energy transfer participates in EL process. At the current density of 10.0 and 80.0 mA/cm 2 , 2% and 3% doped devices show the highest EL efficiency, respectively. From analysis of the EL spectra and the EL efficiency-current density characteristics, we found that the EL efficiency is manipulated by Foerster energy transfer efficiency at high current density. So we suggest that the dominant luminescent mechanism changes gradually from carrier trapping to Foerster energy transfer with increasing current density. Moreover, the conversion of dominant EL mechanism was suspected to be partly responsible for the EL efficiency roll-off because of the lower EL quantum efficiency of Foerster energy transfer compared with carrier trapping

  16. Current state of low energy EB devices and its application technology

    International Nuclear Information System (INIS)

    Kinoshita, Shinobu

    2000-01-01

    This paper introduced the current state of low energy type EB (electron beam) devices with an acceleration voltage of 300 kV or below and specific application examples. As for EB devices, it introduced the ultra-compact new EB device (microbeam LV), experimental devices, and the pilot/production devices which have been recently developed by the manufacturer to which the author belongs. As the applications of low energy EB devices, it specifically introduced curing, graft polymerization, crosslinking, and sterilization/disinfection with soft electrons: (1) examples of EB curing; antistatic agents in antibacterial/antifungal property imparting processing, hard coat, printing and topcoat, high gloss/pattern transfer processing, and metal vapor deposition film, (2) example of graft polymerization; barrier imparting films, and (3) examples of crosslinking; shrinking films/tubes and foamed sheets. (A.O.)

  17. Method of controlling illumination device based on current-voltage model

    DEFF Research Database (Denmark)

    2013-01-01

    The present invention relates to an illumination device comprising a number of LEDs, means for receiving an input signal, means for generating an activation signal for at least one of the LEDs based on the input signal. The illumination device comprises further means for obtaining the voltage...... and the colorimetric properties of said light emitted by LED. The present invention relates also to a method of controlling and a meted of calibrating such illumination device....... across and current through the LED and the means for generating the activation signal is adapted to generate the activating signal based on the voltage, the current and a current- voltage model related to LED. The current-voltage model defines a relationship between the current, the voltage...

  18. Circuit for current measures from ionization chambers

    International Nuclear Information System (INIS)

    Mello, F.L.V. de; Oliveira, A.H. de; Rezende, R.S.

    1992-01-01

    The design and the specifications of an ammeters of low cost for small current, IOE-14 Ampere, from ionization chambers or others transducers used in nuclear instrumentation are described. Special attention is given to the integrated electronic components, available in the brazilian market. (C.G.C.)

  19. Gain stabilization circuit of measuring devices with photomultipliers

    International Nuclear Information System (INIS)

    Seda, J.; Sabol, J.

    1974-01-01

    A circuit is designed for the stabilization of the gain of measuring devices with photomultipliers, suitable especially for the stabilization of scintillation detection systems, in which the correction signal is applied to the photomultiplier grid placed between the photocathode and the first dynode. (J.K.)

  20. PEPR and other CRT scanning and measuring devices

    Energy Technology Data Exchange (ETDEWEB)

    Pless, Irwin A. (Massachusetts Institute of Technology, Cambridge, MA 02139 (United States))

    1994-07-01

    This talk concentrates on the brief history of the PEPR (Precision Encoding and Pattern Recognition) scanning and measuring device.I restrict this reminiscence to just a short history of PEPR, the other various CRT scanners, a short summary of the capabilities of this scanner and some fond memories. ((orig.))

  1. Electron density measurement in an evolving plasma. Experimental devices

    International Nuclear Information System (INIS)

    Consoli, Terenzio; Dagai, Michel

    1960-01-01

    The experimental devices described here allow the electron density measurements in the 10 16 e/m 3 to 10 20 e/m 3 interval. Reprint of a paper published in Comptes rendus des seances de l'Academie des Sciences, t. 250, p. 1223-1225, sitting of 15 February 1960 [fr

  2. PEPR and other CRT scanning and measuring devices

    International Nuclear Information System (INIS)

    Pless, Irwin A.

    1994-01-01

    This talk concentrates on the brief history of the PEPR (Precision Encoding and Pattern Recognition) scanning and measuring device.I restrict this reminiscence to just a short history of PEPR, the other various CRT scanners, a short summary of the capabilities of this scanner and some fond memories. ((orig.))

  3. 21 CFR 886.1450 - Corneal radius measuring device.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Corneal radius measuring device. 886.1450 Section 886.1450 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES... corneal size by superimposing the image of the cornea on a scale at the focal length of the lens of a...

  4. Wave measurement in severe ocean currents

    Digital Repository Service at National Institute of Oceanography (India)

    Diwan, S.G.; Suryavanshi, A.K.; Nayak, B.U.

    The measurement of ocean waves has been of particular interest, as wave data and understanding of wave phenomena are essential to ocean engineering, coastal engineering and to many marine operations. The National Institute of Oceanography, Goa...

  5. Bidirectional current triggering in planar devices based on serially connected VO2 thin films using 965 nm laser diode.

    Science.gov (United States)

    Kim, Jihoon; Park, Kyongsoo; Kim, Bong-Jun; Lee, Yong Wook

    2016-08-08

    By incorporating a 965 nm laser diode, the bidirectional current triggering of up to 30 mA was demonstrated in a two-terminal planar device based on serially connected vanadium dioxide (VO2) thin films grown by pulsed laser deposition. The bidirectional current triggering was realized by using the focused beams of laser pulses through the photo-thermally induced phase transition of VO2. The transient responses of laser-triggered currents were also investigated when laser pulses excited the device at a variety of pulse widths and repetition rates of up to 4.0 Hz. A switching contrast between off- and on-state currents was obtained as ~8333, and rising and falling times were measured as ~39 and ~29 ms, respectively, for 50 ms laser pulses.

  6. Assembly for electrical conductivity measurements in the piston cylinder device

    Science.gov (United States)

    Watson, Heather Christine [Dublin, CA; Roberts, Jeffrey James [Livermore, CA

    2012-06-05

    An assembly apparatus for measurement of electrical conductivity or other properties of a sample in a piston cylinder device wherein pressure and heat are applied to the sample by the piston cylinder device. The assembly apparatus includes a body, a first electrode in the body, the first electrode operatively connected to the sample, a first electrical conductor connected to the first electrode, a washer constructed of a hard conducting material, the washer surrounding the first electrical conductor in the body, a second electrode in the body, the second electrode operatively connected to the sample, and a second electrical conductor connected to the second electrode.

  7. On a mechanism of switching off low-hybrid run away currents in tokamak devices

    International Nuclear Information System (INIS)

    Budnikov, V.N.; Esipov, L.A.; Irzak, M.A.

    1990-01-01

    The problem of the generation of low-hybrid run-away currents (LR) in tokamak devices is described. The mechanism of switching off LRCs is considered. Qualitative representation of the density limit, the transitions of which stops the generation of currents, is given

  8. Device for plasma confinement and heating by high currents and nonclassical plasma transport properties

    Science.gov (United States)

    Coppi, B.; Montgomery, D.B.

    1973-12-11

    A toroidal plasma containment device having means for inducing high total plasma currents and current densities and at the same time emhanced plasma heating, strong magnetic confinement, high energy density containment, magnetic modulation, microwaveinduced heating, and diagnostic accessibility is described. (Official Gazette)

  9. Modelling band-to-band tunneling current in InP-based heterostructure photonic devices

    NARCIS (Netherlands)

    van Engelen, J.P.; Shen, L.; van der Tol, J.J.G.M.; Smit, M.K.; Kockaert, P.; Emplit, P.; Gorza, S.-P.; Massar, S.

    2015-01-01

    Some semiconductor photonic devices show large discontinuities in the band structure. Short tunnel paths caused by this band structure may lead to an excessive tunneling current, especially in highly doped layers. Modelling of this tunnelling current is therefore important when designing photonic

  10. Digital DC beam current measurement on SSRF storage ring

    International Nuclear Information System (INIS)

    Xiong Liang; Yin Chongxian; Liu Ming; Chen Jianfeng

    2011-01-01

    Both DC current transformer (DCCT) and integrating current transformer (ICT) can be used in DC beam current measurement. The ICT has strong capability of resisting electromagnetic interference, but its measurement accuracy cannot satisfy the DC beam current measurement requirement when using traditional high speed A/D. With high resolution A/D and equivalent sampling system, DC beam current measuring system based on ICT can reach high accuracy compared with DCCT system. In this paper, the ICT-based DC beam current measurement, equivalent sampling method and testing results at Shanghai Synchrotron Radiation Facility(SSRF) is described. (authors)

  11. A device for automated phase space measurement of ion beams

    International Nuclear Information System (INIS)

    Lukas, J.; Priller, A.; Steier, P.

    2007-01-01

    Equipment for automated phase-space measurements was developed at the VERA Lab. The measurement of the beam's intensity distribution, as well as its relative position with respect to the reference orbit is performed at two locations along the beam line. The device basically consists of moveable slits and a beam profile monitor, which are both coordinated and controlled by an embedded controller. The operating system of the controller is based on Linux with real-time extension. It controls the movement of the slits and records the data synchronized to the movement of the beam profile monitor. The data is sent via TCP/IP to the data acquisition system of VERA where visualization takes place. The duration of one phase space measurement is less than 10 s, which allows for using the device during routine beam tuning

  12. Single-Image Distance Measurement by a Smart Mobile Device.

    Science.gov (United States)

    Chen, Shangwen; Fang, Xianyong; Shen, Jianbing; Wang, Linbo; Shao, Ling

    2017-12-01

    Existing distance measurement methods either require multiple images and special photographing poses or only measure the height with a special view configuration. We propose a novel image-based method that can measure various types of distance from single image captured by a smart mobile device. The embedded accelerometer is used to determine the view orientation of the device. Consequently, pixels can be back-projected to the ground, thanks to the efficient calibration method using two known distances. Then the distance in pixel is transformed to a real distance in centimeter with a linear model parameterized by the magnification ratio. Various types of distance specified in the image can be computed accordingly. Experimental results demonstrate the effectiveness of the proposed method.

  13. A TECHNIQUE OF MEASURING OF RESISTANCE OF A GROUNDING DEVICE

    Directory of Open Access Journals (Sweden)

    I.V. Nizhevskyi

    2016-06-01

    Full Text Available Introduction. Measurement of resistance of the grounding device (GD by means of a three-electrode system. This requires not only the right choice of installation locations of measuring electrodes, but also the determination of the point of zero potential. Implementation of these requirements quite time-consuming, and in some cases impossible. Aim. Develop a new technique for measuring the electrical resistance of the GD. Task. The method of measuring the resistance of the GD with the help of a three-electrode setup is necessary to exclude the determination of the point of zero potential. Method. Mathematical modeling and calculation engine. Results. A three-electrode system for measuring the resistance of grounding devices (GD for various purposes is considered. On the basis of Maxwell equations a theoretical substantiation of a new technique for measuring the resistance of any GD of any construction in random soil structure has been proposed. An equation system of the sixth order has been obtained, its solution makes it possible to measure its own mutual resistance in the three-electrode installation with sufficiently high accuracy. Peculiarities of drawing up a calculation scheme of substitution of a three-electrode installation with lumped parameters: self and mutual impedance. Use of the principle of reciprocity eliminates the need of finding a point of zero potential which is a rather difficult task. The technique allows to minimize the spacing of measuring electrodes outside the GD, which substantially reduces the length of wiring of the measurement circuit and increases the «signal-to-interference» ratio and also removes the restrictions on the development of the territory outside the GD being tested. Conclusion. The procedure allows to evaluate the self and mutual impedance grounding all the electrodes in a three-electrode measuring installation of the grounding resistance of the device without finding the point of zero potential.

  14. Improving the accuracy of smart devices to measure noise exposure.

    Science.gov (United States)

    Roberts, Benjamin; Kardous, Chucri; Neitzel, Richard

    2016-11-01

    Occupational noise exposure is one of the most frequent hazards present in the workplace; up to 22 million workers have potentially hazardous noise exposures in the U.S. As a result, noise-induced hearing loss is one of the most common occupational injuries in the U.S. Workers in manufacturing, construction, and the military are at the highest risk for hearing loss. Despite the large number of people exposed to high levels of noise at work, many occupations have not been adequately evaluated for noise exposure. The objective of this experiment was to investigate whether or not iOS smartphones and other smart devices (Apple iPhones and iPods) could be used as reliable instruments to measure noise exposures. For this experiment three different types of microphones were tested with a single model of iPod and three generations of iPhones: the internal microphones on the device, a low-end lapel microphone, and a high-end lapel microphone marketed as being compliant with the International Electrotechnical Commission's (IEC) standard for a Class 2-microphone. All possible combinations of microphones and noise measurement applications were tested in a controlled environment using several different levels of pink noise ranging from 60-100 dBA. Results were compared to simultaneous measurements made using a Type 1 sound level measurement system. Analysis of variance and Tukey's honest significant difference (HSD) test were used to determine if the results differed by microphone or noise measurement application. Levels measured with external microphones combined with certain noise measurement applications did not differ significantly from levels measured with the Type 1 sound measurement system. Results showed that it may be possible to use iOS smartphones and smart devices, with specific combinations of measurement applications and calibrated external microphones, to collect reliable, occupational noise exposure data under certain conditions and within the limitations of the

  15. Real-time precision measuring device of tree diameter growth

    Science.gov (United States)

    Guo, Mingming; Chen, Aijun; Li, Dongsheng; Liu, Nan; Yao, Jingyuan

    2016-01-01

    DBH(diameter at breast height) is an important factor to reflect of the quality of plant growth, also an important parameter indispensable in forest resources inventory and forest carbon sink, the accurate measurement of DBH or not is directly related to the research of forest resources inventory and forest carbon sink. In this paper, the principle and the mathematical model of DBH measurement device were introduced, the fixture measuring device and the hardware circuit for this tree diameter were designed, the measurement software programs were compiled, and the precision measuring device of tree diameter growth was developed. Some experiments with Australia fir were conducted. Based on experiment data, the correlations among the DBH variation of Australian fir, the environment temperature, air humility and PAR(photosynthetically active radiation) were obtained. The effects of environmental parameters (environment temperature, air humility and PAR) on tree diameter were analyzed. Experimental results show that there is a positive correlation between DBH variation of Australian fir and environment temperature, a negative correlation between DBH variation of Australian fir and air humility , so is PAR.

  16. Current measurement method for characterization of fast switching power semiconductors with Silicon Steel Current Transformer

    DEFF Research Database (Denmark)

    Li, Helong; Beczkowski, Szymon; Munk-Nielsen, Stig

    2015-01-01

    This paper proposes a novel current measurement method with Silicon Steel Current Transformer (SSCT) for the characterization of fast switching power semiconductors. First, the existing current sensors for characterization of fast switching power semiconductors are experimentally evaluated...

  17. Proof of Concept: Design and Initial Evaluation of a Device to Measure Gastrointestinal Transit Time.

    Science.gov (United States)

    Wagner, Robert H; Savir-Baruch, Bital; Halama, James R; Venu, Mukund; Gabriel, Medhat S; Bova, Davide

    2017-09-01

    Chronic constipation and gastrointestinal motility disorders constitute a large part of a gastroenterology practice and have a significant impact on a patient's quality of life and lifestyle. In most cases, medications are prescribed to alleviate symptoms without there being an objective measurement of response. Commonly used investigations of gastrointestinal transit times are currently limited to radiopaque markers or electronic capsules. Repeated use of these techniques is limited because of the radiation exposure and the significant cost of the devices. We present the proof of concept for a new device to measure gastrointestinal transit time using commonly available and inexpensive materials with only a small amount of radiotracer. Methods: We assembled gelatin capsules containing a 67 Ga-citrate-radiolabeled grain of rice embedded in paraffin for use as a point-source transit device. It was tested for stability in vitro and subsequently was given orally to 4 healthy volunteers and 10 patients with constipation or diarrhea. Imaging was performed at regular intervals until the device was excreted. Results: The device remained intact and visible as a point source in all subjects until excretion. When used along with a diary of bowel movement times and dates, the device could determine the total transit time. The device could be visualized either alone or in combination with a barium small-bowel follow-through study or a gastric emptying study. Conclusion: The use of a point-source transit device for the determination of gastrointestinal transit time is a feasible alternative to other methods. The device is inexpensive and easy to assemble, requires only a small amount of radiotracer, and remains inert throughout the gastrointestinal tract, allowing for accurate determination of gastrointestinal transit time. Further investigation of the device is required to establish optimum imaging parameters and reference values. Measurements of gastrointestinal transit time

  18. Enhancement of superconducting critical current by injection of quasiparticles in superconductor semiconductor devices

    DEFF Research Database (Denmark)

    Kutchinsky, Jonatan; Taboryski, Rafael Jozef; Sørensen, C. B.

    2000-01-01

    We report new measurements on 3-terminal superconductor semiconductor injection devices, demonstrating enhancement of the supercurrent by injection from a superconducting injector electrode. Two other electrodes were used as detectors. Applying a small voltage to the injector, reduced the maximum...

  19. Current State and Future Perspectives of Energy Sources for Totally Implantable Cardiac Devices.

    Science.gov (United States)

    Bleszynski, Peter A; Luc, Jessica G Y; Schade, Peter; PhilLips, Steven J; Tchantchaleishvili, Vakhtang

    There is a large population of patients with end-stage congestive heart failure who cannot be treated by means of conventional cardiac surgery, cardiac transplantation, or chronic catecholamine infusions. Implantable cardiac devices, many designated as destination therapy, have revolutionized patient care and outcomes, although infection and complications related to external power sources or routine battery exchange remain a substantial risk. Complications from repeat battery replacement, power failure, and infections ultimately endanger the original objectives of implantable biomedical device therapy - eliminating the intended patient autonomy, affecting patient quality of life and survival. We sought to review the limitations of current cardiac biomedical device energy sources and discuss the current state and trends of future potential energy sources in pursuit of a lifelong fully implantable biomedical device.

  20. UNIVERSAL AUTO-CALIBRATION FOR A RAPID BATTERY IMPEDANCE SPECTRUM MEASUREMENT DEVICE

    Energy Technology Data Exchange (ETDEWEB)

    Jon P. Christophersen; John L. Morrison; William H. Morrison

    2014-03-01

    Electrochemical impedance spectroscopy has been shown to be a valuable tool for diagnostics and prognostics of energy storage devices such as batteries and ultra-capacitors. Although measurements have been typically confined to laboratory environments, rapid impedance spectrum measurement techniques have been developed for on-line, embedded applications as well. The prototype hardware for the rapid technique has been validated using lithium-ion batteries, but issues with calibration had also been identified. A new, universal automatic calibration technique was developed to address the identified issues while also enabling a more simplified approach. A single, broad-frequency range is used to calibrate the system and then scaled to the actual range and conditions used when measuring a device under test. The range used for calibration must be broad relative to the expected measurement conditions for the scaling to be successful. Validation studies were performed by comparing the universal calibration approach with data acquired from targeted calibration ranges based on the expected range of performance for the device under test. First, a mid-level shunt range was used for calibration and used to measure devices with lower and higher impedance. Next, a high excitation current level was used for calibration, followed by measurements using lower currents. Finally, calibration was performed over a wide frequency range and used to measure test articles with a lower set of frequencies. In all cases, the universal calibration approach compared very well with results acquired following a targeted calibration. Additionally, the shunts used for the automated calibration technique were successfully characterized such that the rapid impedance measurements compare very well with laboratory-scale measurements. These data indicate that the universal approach can be successfully used for onboard rapid impedance spectra measurements for a broad set of test devices and range of

  1. Radiation detection and measurement concepts, methods and devices

    CERN Document Server

    McGregor, Douglas

    2019-01-01

    This text on radiation detection and measurement is a response to numerous requests expressed by students at various universities, in which the most popularly used books do not provide adequate background material, nor explain matters in understandable terms. This work provides a modern overview of radiation detection devices and radiation measurement methods. The topics selected in the book have been selected on the basis of the author’s many years of experience designing radiation detectors and teaching radiation detection and measurement in a classroom environment.

  2. Device for the integral measurement of ionizing radiations

    International Nuclear Information System (INIS)

    Micheron, Francois.

    1980-01-01

    This invention relates to devices for the integral determination of ionizing radiations, particularly to the construction of a portable dosemeter. Portable measuring instruments have been suggested in the past, particularly dosemeters in which the discharge of a capacitor under the action of ionizing radiations is measured. Since the charge of a capacitor is not stable owing to dielectric imperfections, these measuring instruments have to be recalibrated at frequent intervals. To overcome this drawback, the invention suggests using the discharge of an electret, electrically charged to a pre-set initial value, under the action of ionizing radiations, as the transducer means of a dosemeter used in conjunction with display or warning systems [fr

  3. The DECMU: a digital device for delayed analysis of multi-frequency eddy current signals

    International Nuclear Information System (INIS)

    Pigeon, Michel.

    1981-08-01

    A delayed data analysis system has been realized by the CEA and Intercontrole for in-service inspection of steam generators of nuclear plants by multifrequency eddy current testing. This device allows, out of the plant, adjustment during switching of the probes, graph recording and analysis for defect signal qualification. The equipment contains an analog mixing device, as IC3FA multi-frequency appartus, but has in addition a memory allowing data cycling and signal isolation for adjustment or analysis [fr

  4. Eddy Current Flow Measurements in the FFTF

    Energy Technology Data Exchange (ETDEWEB)

    Nielsen, Deborah L. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Polzin, David L. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Omberg, Ronald P. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Makenas, Bruce J. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States)

    2017-02-02

    The Fast Flux Test Facility (FFTF) is the most recent liquid metal reactor (LMR) to be designed, constructed, and operated by the U.S. Department of Energy (DOE). The 400-MWt sodium-cooled, fast-neutron flux reactor plant was designed for irradiation testing of nuclear reactor fuels and materials for liquid metal fast breeder reactors. Following shut down of the Clinch River Breeder Reactor Plant (CRBRP) project in 1983, FFTF continued to play a key role in providing a test bed for demonstrating performance of advanced fuel designs and demonstrating operation, maintenance, and safety of advanced liquid metal reactors. The FFTF Program provides valuable information for potential follow-on reactor projects in the areas of plant system and component design, component fabrication, fuel design and performance, prototype testing, site construction, and reactor control and operations. This report provides HEDL-TC-1344, “ECFM Flow Measurements in the FFTF Using Phase-Sensitive Detectors”, March 1979.

  5. Numerical simulation of current-free double layers created in a helicon plasma device

    Science.gov (United States)

    Rao, Sathyanarayan; Singh, Nagendra

    2012-09-01

    Two-dimensional simulations reveal that when radially confined source plasma with magnetized electrons and unmagnetized ions expands into diverging magnetic field B, a current-free double layer (CFDL) embedded in a conical density structure forms, as experimentally measured in the Australian helicon plasma device (HPD). The magnetized electrons follow the diverging B while the unmagnetized ions tend to flow directly downstream of the source, resulting in a radial electric field (E⊥) structure, which couples the ion and electron flows. Ions are transversely (radially) accelerated by E⊥ on the high potential side of the double layer in the CFDL. The accelerated ions are trapped near the conical surface, where E⊥ reverses direction. The potential structure of the CFDL is U-shaped and the plasma density is enhanced on the conical surface. The plasma density is severely depleted downstream of the parallel potential drop (φ||o) in the CFDL; the density depletion and the potential drop are related by quasi-neutrality condition, including the divergence in the magnetic field and in the plasma flow in the conical structure. The potential and density structures, the CFDL spatial size, its electric field strengths and the electron and ion velocities and energy distributions in the CFDL are found to be in good agreements with those measured in the Australian experiment. The applicability of our results to measured axial potential profiles in magnetic nozzle experiments in HPDs is discussed.

  6. Head-Impact–Measurement Devices: A Systematic Review

    Science.gov (United States)

    O'Connor, Kathryn L.; Rowson, Steven; Duma, Stefan M.; Broglio, Steven P.

    2017-01-01

    Context: With an estimated 3.8 million sport- and recreation-related concussions occurring annually, targeted prevention and diagnostic methods are needed. Biomechanical analysis of head impacts may provide quantitative information that can inform both prevention and diagnostic strategies. Objective: To assess available head-impact devices and their clinical utility. Data Sources: We performed a systematic search of the electronic database PubMed for peer-reviewed publications, using the following phrases: accelerometer and concussion, head impact telemetry, head impacts and concussion and sensor, head impacts and sensor, impact sensor and concussion, linear acceleration and concussion, rotational acceleration and concussion, and xpatch concussion. In addition to the literature review, a Google search for head impact monitor and concussion monitor yielded 15 more devices. Study Selection: Included studies were performed in vivo, used commercially available devices, and focused on sport-related concussion. Data Extraction: One author reviewed the title and abstract of each study for inclusion and exclusion criteria and then reviewed each full-text article to confirm inclusion criteria. Controversial articles were reviewed by all authors to reach consensus. Data Synthesis: In total, 61 peer-reviewed articles involving 4 head-impact devices were included. Participants in boxing, football, ice hockey, soccer, or snow sports ranged in age from 6 to 24 years; 18% (n = 11) of the studies included female athletes. The Head Impact Telemetry System was the most widely used device (n = 53). Fourteen additional commercially available devices were presented. Conclusions: Measurements collected by impact monitors provided real-time data to estimate player exposure but did not have the requisite sensitivity to concussion. Proper interpretation of previously reported head-impact kinematics across age, sport, and position may inform future research and enable staff clinicians

  7. Surface photovoltage measurements and finite element modeling of SAW devices.

    Energy Technology Data Exchange (ETDEWEB)

    Donnelly, Christine

    2012-03-01

    Over the course of a Summer 2011 internship with the MEMS department of Sandia National Laboratories, work was completed on two major projects. The first and main project of the summer involved taking surface photovoltage measurements for silicon samples, and using these measurements to determine surface recombination velocities and minority carrier diffusion lengths of the materials. The SPV method was used to fill gaps in the knowledge of material parameters that had not been determined successfully by other characterization methods. The second project involved creating a 2D finite element model of a surface acoustic wave device. A basic form of the model with the expected impedance response curve was completed, and the model is ready to be further developed for analysis of MEMS photonic resonator devices.

  8. A handheld optical device for skin profile measurement

    Science.gov (United States)

    Sun, Jiuai; Liu, Xiaojin

    2018-04-01

    This paper describes a portable optical scanning device designed for skin surface measurement on both colour and 3D geometry through a relative easy and cost effective multiple light source photometric stereo method. The validation of colour recovered had been verified through its application on skin lesion segmentation in our early work. This paper focuses on the reconstructed topographic data which are subject to further evaluation and advancement. The evaluation work takes the skin in vitro as an application scenario and compares the experimental result to that obtained by using a commercial product. The experiments show that this handheld device can measure the skin profile significantly closer to that of the ground truth and have the additional function of skin colour recovery.

  9. Post market surveillance in the german medical device sector - current state and future perspectives.

    Science.gov (United States)

    Zippel, Claus; Bohnet-Joschko, Sabine

    2017-08-01

    Medical devices play a central role in the diagnosis and treatment of diseases but also bring the potential for adverse events, hazards or malfunction with serious consequences for patients and users. Medical device manufacturers are therefore required by law to monitor the performance of medical devices that have been approved by the competent authorities (post market surveillance). Conducting a nationwide online-survey in the German medical device sector in Q2/2014 in order to explore the current status of the use of post market instruments we obtained a total of 118 complete data sets, for a return rate of 36%. The survey included manufacturers of different sizes, producing medical devices of all risk classes. The post market instruments most frequently reported covered the fields of production monitoring and quality management as well as literature observation, regulatory vigilance systems, customer knowledge management and market observation while Post Market Clinical Follow-up and health services research were being used less for product monitoring. We found significant differences between the different risk classes of medical devices produced and the intensity of use of post market instruments. Differences between company size and the intensity of instruments used were hardly detected. Results may well contribute to the development of device monitoring which is a crucial element of the policy and regulatory system to identify device-related safety issues. Copyright © 2017 Elsevier B.V. All rights reserved.

  10. Development of signal acquisition device of rotating coil measurement system

    International Nuclear Information System (INIS)

    Zhou Jianxin; Li Li; Kang Wen; Deng Chengdong; Yin Baogui; Fu Shinian

    2013-01-01

    A new rotating coil magnetic measurement system using the technical solution of the combination of a dynamic signal acquisition card and software with specific functions was developed. The acquisition device of the system successfully implemented the function of the PDI-5025 integrator. The sampling rate, the range, the accuracy and the flexibility of the system were improved. The development program of signal acquisition equipment, the realization of the acquisition function and the reliability and stability of the system were introduced. (authors)

  11. Design of Gear Churning Power Loss Measurement Device

    OpenAIRE

    Wang Bin; Zhou Ya Jie; Wang Ping

    2017-01-01

    To explore the impacts of gear churning power losses, a research was conducted to achieve the internal causes of power losses of churning gear by designing a gear churning power losses measurement device. The gear churning power losses could be influenced by different gear modules, the number of teeth and the axial position of gear. Finally, the impacts of gear churning power losses were discussed by comparing experimental data and theoretical data.

  12. Dark properties and transient current response of Si0.95Ge0.05 n+p devices

    International Nuclear Information System (INIS)

    Ruzin, Arie; Marunko, S.; Abrosimov, N.V.; Riemann, H.

    2004-01-01

    In this study we present the dark properties of 'pin' devices fabricated with Czochralski grown Si 0.95 Ge 0.05 bulk single crystals. The growth of such material is most challenging because of the constitutional supercooling effect. The potential advantages of Si 1-x Ge x to be used for X- and gamma-ray detection applications are overviewed. At room temperature the generation current in the devices is too high for spectroscopy applications, but enables transient current technique (TCT) measurements. The current however drops significantly with moderate cooling. The effective majority carrier concentration is shown to be ∼2x10 14 cm -3 , and hole mobility ∼320 cm 2 /V s

  13. Opto-mechanical design of small infrared cloud measuring device

    Science.gov (United States)

    Zhang, Jiao; Yu, Xun; Tao, Yu; Jiang, Xu

    2018-01-01

    In order to make small infrared cloud measuring device can be well in a wide temperature range and day-night environment, a design idea using catadioptric infrared panoramic imaging optical system and simple mechanical structure for realizing observation clode under all-weather conditions was proposed. Firstly, the optical system of cloud measuring device was designed. An easy-to-use numerical method was proposed to acquire the profile of a catadioptric mirror, which brought the property of equidistance projection and played the most important role in a catadioptric panoramic lens. Secondly, the mechanical structure was studied in detail. Overcoming the limitations of traditional primary mirror support structure, integrative design was used for refractor and mirror support structure. Lastly, temperature adaptability and modes of the mirror support structure were analyzed. Results show that the observation range of the cloud measuring device is wide and the structure is simple, the fundamental frequency of the structure is greater than 100 Hz, the surface precision of the system reflector reaches PV of λ/10 and RMS of λ/40under the load of temperature range - 40 60°C, it can meet the needs of existing meteorological observation.

  14. The making of pressure measurement device on heating-02 based realtime

    International Nuclear Information System (INIS)

    Giarno; Kussigit Santosa; Agus Nur Rachman; G B Heru K

    2013-01-01

    In order to modify the installation strand BETA Test Section Test integrated with heating-02 into a closed loop, it would require an additional system that can measure pressure changes in the closed-loop system. By making the measurement device to test the system pressure at the heating-expected 02 researchers can monitor the pressure changes that occur in the system. The pressure gauge device fabrication using manufacturing simulation methodology, the preparation of the hardware and software and test functions. Manufacturing simulation using measuring devices HIOKI DC current source Signal Source, preparation of pressure measurement devices require hardware such as pressure transducers, NI cRIO-9074, NI 9203 analog module, Computer and software LabVIEW 2011 as programming. In the test process function method is used to provide flow simulation module that is connected to the 9203 NI NI cRIO-9074. Current provision tailored to the specifics pressure transducer is 4 mA s/d 20 mA. Based on the test results obtained function value of the lowest current is 4.00 mA = 0.001 bar, and the highest current value of 20.00 mA = 4995 bar. From the results of calculations using the linear equations obtained correlation coefficient (R 2 ) of 0.999, so it is evident that the pressure changes in LabVIEW is affected by changes in flow. The results obtained from this activity is a device that can measure the pressure in the heating-02 test. (author)

  15. MEASURING RESULTS NUMERAL TREATMENT OF IMPULSIVE CURRENTS BY MEANS OF ROGOVSKY BELT APPLICATION

    Directory of Open Access Journals (Sweden)

    U. Batygin

    2009-01-01

    Full Text Available The technique of numerical processing of measurement results of pulse currents by means of Rogovsky belt application is offered in the given work. It is shown that at measurement of signals by digital oscillographs and further numerical transformation of target signals, the possibilities of Rogovsky belt without the application of additional devices that in turn allows to define parameters of pulse currents with any peak-time characteristics essentially expand.

  16. Integrated-optic current sensors with a multimode interference waveguide device.

    Science.gov (United States)

    Kim, Sung-Moon; Chu, Woo-Sung; Kim, Sang-Guk; Oh, Min-Cheol

    2016-04-04

    Optical current sensors based on polarization-rotated reflection interferometry are demonstrated using polymeric integrated optics and various functional optical waveguide devices. Interferometric sensors normally require bias feedback control for maintaining the operating point, which increases the cost. In order to resolve this constraint of feedback control, a multimode interference (MMI) waveguide device is integrated onto the current-sensor optical chip in this work. From the multiple outputs of the MMI, a 90° phase-shifted transfer function is obtained. Using passive quadrature demodulation, we demonstrate that the sensor could maintain the output signal regardless of the drift in the operating bias-point.

  17. Development of net-current free heliotron plasmas in the Large Helical Device

    International Nuclear Information System (INIS)

    Komori, A.; Yamada, H.; Kaneko, O.; Kawahata, K.; Mutoh, T.; Ohyabu, N.; Imagawa, S.; Ida, K.; Nagayama, Y.; Shimozuma, T.; Watanabe, K.Y.; Mito, T.; Kobayashi, M.; Nagaoka, K.; Sakamoto, R.; Ohdachi, S.; Sakakibara, S.; Ashikawa, N.; Igami, H.; Kasahara, H.; Kubo, S.; Kumazawa, R.; Nishiura, M.; Masuzaki, S.; Tanaka, K.; Toi, K.; Yoshinuma, M.; Narushima, Y.; Tamura, N.; Saito, K.; Seki, T.; Sudo, S.; Tanaka, H.; Tokuzawa, T.; Yanagi, N.; Yokoyama, M.; Yoshimura, Y.; Akiyama, T.; Chikaraishi, H.; Emoto, M.; Funaba, H.; Goncharov, P.; Goto, M.; Ichiguchi, K.; Ido, T.; Ikeda, K.; Yoshida, N.; Inagaki, S.; Idei, H.; Feng, Y.; Weller, A.; Fukuda, T.; Mitarai, O.; Murakami, S.; Nakamura, Y.; Hino, T.; Ohno, N.; Okamura, T.; Iio, S.; Chowdhuri, M.; Ezumi, N.; Garcia, L.; Ichimura, M.; Irie, M.; Isayama, Akihiko; Iwamae, Atsushi; Takenaga, Hidenobu; Urano, Hajime

    2008-10-01

    Remarkable progress in the physical parameters of net-current free plasmas has been made in the Large Helical Device (LHD) since the last Fusion Energy Conference in Chengdu, 2006 (O. Motojima et al., Nucl. Fusion 47 (2007) S668). The beta value reached 5 % and a high beta state beyond 4.5% from the diamagnetic measurement has been maintained for longer than 100 times the energy confinement time. The density and temperature regimes also have been extended. The central density has exceeded 1.0x10 21 m -3 due to the formation of an Internal Diffusion Barrier (IDB). The ion temperature has reached 6.8 keV at the density of 2x10 19 m -3 , which is associated with the suppression of ion heat conduction loss. Although these parameters have been obtained in separated discharges, each fusion-reactor relevant parameter has elucidated the potential of net-current free heliotron plasmas. Diversified studies in recent LHD experiments are reviewed in this paper. (author)

  18. Speckle interferometry application for erosion measurements in fusion devices

    Energy Technology Data Exchange (ETDEWEB)

    Gauthier, E.; Roupillard, R. [Association Euratom-CEA Cadarache, 13 - Saint-Paul-lez-Durance (France). Dept. de Recherches sur la Fusion Controlee

    2003-07-01

    In order to measure erosion/redeposition in fusion devices, a new diagnostic based on speckle interferometry is investigated. First experiments performed on carbon fibre composite (CFC) materials have shown that this technique is able to measure a modification of the surface in the range of 1 {mu}m. Further experiments have been performed on different materials using a second wavelength in order to carry out 3-dimensional measurements of the surface and to increase the dynamic range of the depth measurement. A diagnostic, based on two-wavelength TV-holography to measure in situ erosion/redeposition during long duration discharges on the CIEL limiter in Tore Supra, is under development at CEA Cadarache. (authors)

  19. Using Smart Devices to Measure Intermittent Noise in the Workplace

    Directory of Open Access Journals (Sweden)

    Benjamin Roberts

    2017-01-01

    Full Text Available Purpose: To determine the accuracy of smart devices (iPods to measure intermittent noise and integrate a noise dose in the workplace. Materials and Methods: In experiment 1, four iPods were each paired with a Larson Davis Spark dosimeter and exposed to randomly fluctuating pink noise in a reverberant sound chamber. Descriptive statistics and the mean difference between the iPod and its paired dosimeter were calculated for the 1-s data logged measurements. The calculated time weighted average (TWA was also compared between the devices. In experiment 2, 15 maintenance workers and 14 office workers wore an iPod and dosimeter during their work-shift for a maximum of five workdays. A mixed effects linear regression model was used to control for repeated measures and to determine the effect of the device type on the projected 8-h TWA. Results: In experiment 1, a total of 315,306 1-s data logged measurements were made. The interquartile range of the mean difference fell within ±2.0 A-weighted decibels (dBA, which is the standard used by the American National Standards Institute to classify a type 2 sound level meter. The mean difference of the calculated TWA was within ±0.5 dBA except for one outlier. In experiment 2, the results of the mixed effects model found that, on average, iPods measured an 8-h TWA 1.7 dBA higher than their paired dosimeters. Conclusion: This study shows that iPods have the ability to make reasonably accurate noise measurements in the workplace, but they are not as accurate as traditional noise dosimeters.

  20. Nuclear reactor fuel assembly grid measuring method and device

    International Nuclear Information System (INIS)

    Fink, D.J.; Cooper, F.W. Jr.

    1987-01-01

    A device is described for remotely measuring a dimension of a workpiece, comprising: (a) first means for measuring the dimension of the workpiece; (b) second means for limiting the force exerted by the first means against the workpiece; (c) third means connected to the first means for moving the first means in the X, Y, Z axes simultaneously relative to the workpiece; (d) fourth means for limiting the force exerted by the third means moving the first means in the X, Y, Z axes; and (e) fifth means remote from the workpiece for monitoring and controlling the movement of the first means relative to the workpiece

  1. Electrical measuring device for a high temperature reactor

    International Nuclear Information System (INIS)

    Elter, C.; Handel, H.; Schoening, J.; Schmitt, H.

    1982-01-01

    The device for measuring the low or high neutron flux during start-up or at load is accommodated in an armoured guide tube projecting into the floor. A gas-tight capsule is formed as the measuring column with outer dome with a lid solidly connected by a flange to the armoured tube situated on the side wall of the concrete reactor vessel, together with the armoured guide tube. Two shielding shutters prevent the passage of radiation through the armoured tube. (DG) [de

  2. DATA PROCESSING FROM THE MEASURING DEVICE BALLBAR QC20

    Directory of Open Access Journals (Sweden)

    Matúš Košinár

    2014-03-01

    Full Text Available The paper presents an innovative method of data processing from the measurement device – Ballbar QC20W. It was created with a program for data transformation (Visual Basic.NET and it used Fourier transformation. The paper deals with the measuring method of CNC machine tools using Ballbar QC20W. There is an influence between qualitative parameters of machine tools and qualitative parameters of products (tolerances, roughness, etc.. It is very important to hold the stability of qualitative parameters of products as a key factor of production quality. Therefore, is also important to evaluate the accuracy of machine tools and make prediction of possible accuracy.

  3. Establishment and application of standard devices for radioactivity measurement

    International Nuclear Information System (INIS)

    Zhou Changgui; Li Xingyuan; Chen Zigen

    1991-03-01

    In order to establish the radioactivity measurement standards a 4πβ-γ coincidence apparatus and a 4πγ ionization chamber have been installed in the laboratory. The 4πβ-γ coincidence apparatus is for absolute measurement, and its uncertainty is ±(0.3∼5)%. The 4πγ ionization chamber is for working standard, and its uncertainty is ±(1∼5)%. The combination of these devices can meet the quality requirements controlled by National Verification System in the transfer of radioactivity values

  4. Optical effects of shadow masks on short circuit current of organic photovoltaic devices.

    Science.gov (United States)

    Lin, Chi-Feng; Lin, Bing-Hong; Liu, Shun-Wei; Hsu, Wei-Feng; Zhang, Mi; Chiu, Tien-Lung; Wei, Mau-Kuo; Lee, Jiun-Haw

    2012-03-21

    In this paper, we have employed different shadow masks attached on top of organic photovoltaic (OPV) devices to study the optical effects of the former on the short circuit current (J(SC)). To rule out possible lateral electrical conduction and simplify the optical effects inside the device, a small-molecular heterojunction OPV device with a clear donor/acceptor interface was employed with a hole extraction layer exhibiting high resistance intentionally. Careful calibration with a shadow mask was employed. By attaching two layers of opaque masks in combination with a suitable holder design to shield the light from the edges and backside, the value of J(SC) approached that of the dark current, even under 1-sun radiation. With different illumination areas, we found that the photons illuminating the non-active region of the device contributed to 40% of the J(SC) by optical effect within the width of about 1 mm around the active region. When illuminating the non-active area with 12 mm to the active area, a 5.6 times improvement in the J(SC) was observed when the incident angle was 75°. With the introduction of a microstructured film onto the OPV device and an increase in the reflection from the non-active region, a 15% enhancement of the J(SC) compared to the control device was achieved.

  5. Optical sensors for the measurement of electric current and voltage

    Energy Technology Data Exchange (ETDEWEB)

    Rutgers, W R; Hulshof, H J.M.; Laurensse, I J; van der Wey, A H

    1987-01-01

    Optical sensors for the measurement of electrical current and voltage were developed for application in electric power systems. The current sensor, based on the Faraday effect in a monomode glass fiber, and the voltage sensor, based on the transverse Pockels effect in a crystal, are demonstrated in wide-band (10 MHz) interference-free measurements of pulsed currents and impulse voltages.

  6. BEAM-BASED MEASUREMENTS OF PERSISTENT CURRENT DECAY IN RHIC

    International Nuclear Information System (INIS)

    FISCHER, W.; JAIN, A.; TEPIKIAN, S.

    2001-01-01

    The two RHIC rings are equipped with superconducting dipole magnets. At injection, induced persistent currents in these magnets lead to a sextupole component. As the persistent currents decay with time, the horizontal and vertical chromaticities change. From magnet measurements of persistent current decays, chromaticity changes in the machine are estimated and compared with chromaticity measurements

  7. Device intended for measurement of induced trapped charge in insulating materials under electron irradiation in a scanning electron microscope

    International Nuclear Information System (INIS)

    Belkorissat, R; Benramdane, N; Jbara, O; Rondot, S; Hadjadj, A; Belhaj, M

    2013-01-01

    A device for simultaneously measuring two currents (i.e. leakage and displacement currents) induced in insulating materials under electron irradiation has been built. The device, suitably mounted on the sample holder of a scanning electron microscope (SEM), allows a wider investigation of charging and discharging phenomena that take place in any type of insulator during its electron irradiation and to determine accurately the corresponding time constants. The measurement of displacement current is based on the principle of the image charge due to the electrostatic influence phenomena. We are reporting the basic concept and test results of the device that we have built using, among others, the finite element method for its calibration. This last method takes into account the specimen chamber geometry, the geometry of the device and the physical properties of the sample. In order to show the possibilities of the designed device, various applications under different experimental conditions are explored. (paper)

  8. Emission characteristics in solution-processed asymmetric white alternating current field-induced polymer electroluminescent devices

    Science.gov (United States)

    Chen, Yonghua; Xia, Yingdong; Smith, Gregory M.; Gu, Yu; Yang, Chuluo; Carroll, David L.

    2013-01-01

    In this work, the emission characteristics of a blue fluorophor poly(9, 9-dioctylfluorene) (PFO) combined with a red emitting dye: Bis(2-methyl-dibenzo[f,h]quinoxaline)(acetylacetonate)iridium (III) [Ir(MDQ)2(acac)], are examined in two different asymmetric white alternating current field-induced polymer electroluminescent (FIPEL) device structures. The first is a top-contact device in which the triplet transfer is observed resulting in the concentration-dependence of the emission similar to the standard organic light-emitting diode (OLED) structure. The second is a bottom-contact device which, however, exhibits concentration-independence of emission. Specifically, both dye emission and polymer emission are found for the concentrations as high as 10% by weight of the dye in the emitter. We attribute this to the significant different carrier injection characteristics of the two FIPEL devices. Our results suggest a simple and easy way to realize high-quality white emission.

  9. Device for measuring mass of air. Einrichtung zur Luftmassenmessung

    Energy Technology Data Exchange (ETDEWEB)

    Sass, W

    1989-09-28

    In a device for measuring the mass of air, particularly for vehicles with internal combustion engines, with a measurement bridge, in one branch of which an air flow resistance, particularly a hot film sensor, which has air flowing round it, is connected in series with a measuring resistance and in another branch of which a compensation resistance measuring the air temperature is connected in series with a fixed resistor, where the bridge differential voltage is measured in the zero branch of the measuring bridge and the resulting signal is used to control a transistor valve situated in the bridge supply path of a bridge supply source with an emitter connected to the bridge via the transistor base for bridge compensation and where the voltage at the measurement resistance after bridge compensation is evaluated as a measure of the air flow, the invention proposes that the transistor valve should be made as an npn transistor blocking for negative voltage peaks in the bridge supply path. This ensures that for netgative voltage peaks in the supply line, the transistor valve closes temporarily and overheating of the measurement bridge is prevented. Such overheating would lead to measurement of too great air mass flow and therefore to a dangerously too rich fuel/air mixture, for example (instead the negative voltage peaks give a safe temporary lean mixture).

  10. Device for measurement of gas mass flow. Einrichtung zur Gasmassenstrommessung

    Energy Technology Data Exchange (ETDEWEB)

    Sass, W

    1989-09-28

    The invention is concerned with a device for the measurement of gas mass flow, particularly measuring air mass flow for vehicles with internal combustion engines, with a measurement bridge, in one branch of which a gas flow resistance, particularly a hot film sensor, with gas flowing round it, is connected in series with a measurement resistance and in another branch of which a compensation resistance measuring the gas temperature is connected in series with a fixed resistor, where the bridge differential voltage is measured in the zero branch of the measuring bridge and a control parameter is produced from this, in order to control a transistor valve situated in the bridge supply path of a DC voltage source via its control electrode until the bridge is balanced, and where the voltage at the measurement resistance after the bridge is balanced is used as a measure of the gas mass flow. In order to obtain exact results of measurement in spite of relatively high interference noise from the cables, it is proposed that an increased supply DC voltage appreciably decreasing the occurring interference noise from the cables should be produced from a small DC voltage and that the output of the DC/DC voltage converter should be connected to the control electrode of the transistor valve, so that the control parameter for the control electrode is derived from the raised DC supply voltage through reducers depending on the gas flow.

  11. Measurement-device-independent quantum communication with an untrusted source

    Science.gov (United States)

    Xu, Feihu

    2015-07-01

    Measurement-device-independent quantum key distribution (MDI-QKD) can provide enhanced security compared to traditional QKD, and it constitutes an important framework for a quantum network with an untrusted network server. Still, a key assumption in MDI-QKD is that the sources are trusted. We propose here a MDI quantum network with a single untrusted source. We have derived a complete proof of the unconditional security of MDI-QKD with an untrusted source. Using simulations, we have considered various real-life imperfections in its implementation, and the simulation results show that MDI-QKD with an untrusted source provides a key generation rate that is close to the rate of initial MDI-QKD in the asymptotic setting. Our work proves the feasibility of the realization of a quantum network. The network users need only low-cost modulation devices, and they can share both an expensive detector and a complicated laser provided by an untrusted network server.

  12. Instrumentation for Gate Current Noise Measurements on sub-100 nm MOS Transistors

    CERN Document Server

    Gaioni, L; Ratti, L; Re, V; Speziali, V; Traversi, G

    2008-01-01

    This work describes a measuring system that was developed to characterize the gate current noise performances of CMOS devices with minimum feature size in the 100 nm span. These devices play an essential role in the design of present daymixedsignal integrated circuits, because of the advantages associated with the scaling process. The reduction in the gate oxide thickness brought about by CMOS technology downscaling leads to a non-negligible gate current due to direct tunneling phenomena; this current represents a noise source which requires an accurate characterization for optimum analog design. In this paper, two instruments able to perform measurements in two different ranges of gate current values will be discussed. Some of the results of gate current noise characterization will also be presented.

  13. An unheated permeation device for calibrating atmospheric VOC measurements

    Directory of Open Access Journals (Sweden)

    J. Brito

    2011-10-01

    Full Text Available The development of an unpowered permeation device for continuous calibration of in-situ instruments measuring atmospheric volatile organic compounds (VOCs is described. Being lightweight and compact, and containing only negligible amounts of chemicals, the device is especially suited for field use such as on board aircraft. Its speciality is to maintain the permeation process in thermal equilibrium, so that the instantaneous permeation rate can be ascribed to a simple temperature measurement. This equilibrium state is maintained by a combination of three features: (i a thin PTFE membrane as permeation medium which guarantees short stabilization times, (ii a water bath as heat buffer, and (iii a vacuum-panel based insulation, in which features (ii and (iii minimize temperature drifts to ~30 mK h−1 per Kelvin temperature difference to the environment. The respective uncertainty of the permeation rate due to thermal non-equilibrium is kept below 1%. An extensive theory part details the major permeation processes of gases through porous polymers, being Fick's diffusion, Knudsen flow, and viscous flow. Both the measured stabilization time and the measured temperature dependence of the permeation rate independently indicate that the permeation can be described by a viscous flow model, where diffusion of the gas molecules in large pores (having a diameter of >0.05 μm dominates.

  14. Assessment of Current Practice for Tank Testing of Small Marine Energy Devices

    DEFF Research Database (Denmark)

    Kofoed, Jens Peter; Frigaard, Peter

    Discussion Report. Equitable Testing and Evaluation of Marine Energy Extraction Devices in terms of Performance, Cost and Environmental Impact. The report is a contribution by Aalborg University (AAU) to the deliverable on Assessment of current practice for tank testing of small marine energy...

  15. Eddy current testing device for metallic tubes at least locally curved

    International Nuclear Information System (INIS)

    Pigeon, Marcel; Vienot, Claude.

    1975-01-01

    Steam generators, condensers and heat exchangers generally consist of metallic tube bundles, the tubes having a complex geometry. The invention concerns an Eddy current testing device for metallic tubes at least locally curved, operating by translation of a probe inside the tubes [fr

  16. Verification of the short-circuit current making capability of high-voltage switching devices

    NARCIS (Netherlands)

    Smeets, R.P.P.; Linden, van der W.A.

    2001-01-01

    Switching-in of short-circuit current leads to pre-arcing in the switching device. Pre-arcing affects the ability of switchgear to close and latch. In three-phase systems, making is associated with transient voltage phenomena that may have a significant impact on the duration of the pre-arcing

  17. Method for measuring the resistive transition and critical current in superconductors using pulsed current

    International Nuclear Information System (INIS)

    McGinnis, W.C.; Jones, T.E.

    1993-01-01

    A method is described for measuring the intragranular critical current of a granular superconductive material, comprising the steps of: conducting a substantially rectangular electronic pulse through said material so as to conduct a current through said material such that when said intragranular critical current of said material is exceeded, any grains present in said material are in a superconducting state when said current is less than said intragranular critical current, said material having a critical temperature; measuring said current through said material while conducting said pulse; measuring a voltage difference across said material while conducting said pulse; and determining said intragranular critical current through said material by varying said current to discern a current level at which an electrical resistance of said material increases to that of a non-superconducting state as the grains of said material transition from said superconducting to said non-superconducting state

  18. Measurements of ODAK-3K plasma device using plastic track detectors

    International Nuclear Information System (INIS)

    2010-01-01

    In this study, some testing experiments on the fusion researches with a new-constructed plasma focus (PF) device, namely ODAK-3K are reported. The device has a maximal energy input of 3 kJ and is used for both plasma and D D reaction explorations. Experiments with deuterium have shown that peak current of I p eak=39 kA flows between the electrodes at P=11.5 mbar for the operation voltage of V=14 kV. Average total neutron yield is measured around 3.3x10 5 neutrons per shot using CR-39 plastic detectors located opposite the anode inside the PF chamber

  19. The measurement of potential distribution of plasma in MM-4 fusion device

    International Nuclear Information System (INIS)

    Tian Zhongyu; Ming Linzhou; Feng Xiaozhen; Feng Chuntang; Yi Youjun; Wang Jihai; Liu Yihua

    1988-11-01

    Some experimental results of the potential distribution in MM-4 fusion device are presented by measuring the floating potential of probe. The results showed that the distribution of axial potential is asymmetrical, but the radial potential is symmetrical. There are double ion potential wells in the plasma. The depth of the deepest potential well become deeper is the strength of the magnetic field and injection current are increasing. The location of the deepest well is moved towards the device center along with the increasing of injection energy. This is different from others results. The mechanism of causing this distribution in also discussed

  20. Calculation of induced current densities for humans by magnetic fields from electronic article surveillance devices

    Science.gov (United States)

    Gandhi, Om P.; Kang, Gang

    2001-11-01

    This paper illustrates the use of the impedance method to calculate the electric fields and current densities induced in millimetre resolution anatomic models of the human body, namely an adult and 10- and 5-year-old children, for exposure to nonuniform magnetic fields typical of two assumed but representative electronic article surveillance (EAS) devices at 1 and 30 kHz, respectively. The devices assumed for the calculations are a solenoid type magnetic deactivator used at store checkouts and a pass-by panel-type EAS system consisting of two overlapping rectangular current-carrying coils used at entry and exit from a store. The impedance method code is modified to obtain induced current densities averaged over a cross section of 1 cm2 perpendicular to the direction of induced currents. This is done to compare the peak current densities with the limits or the basic restrictions given in the ICNIRP safety guidelines. Because of the stronger magnetic fields at lower heights for both the assumed devices, the peak 1 cm2 area-averaged current densities for the CNS tissues such as the brain and the spinal cord are increasingly larger for smaller models and are the highest for the model of the 5-year-old child. For both the EAS devices, the maximum 1 cm2 area-averaged current densities for the brain of the model of the adult are lower than the ICNIRP safety guideline, but may approach or exceed the ICNIRP basic restrictions for models of 10- and 5-year-old children if sufficiently strong magnetic fields are used.

  1. Calculation of induced current densities for humans by magnetic fields from electronic article surveillance devices.

    Science.gov (United States)

    Gandhi, O P; Kang, G

    2001-11-01

    This paper illustrates the use of the impedance method to calculate the electric fields and current densities induced in millimetre resolution anatomic models of the human body, namely an adult and 10- and 5-year-old children, for exposure to nonuniform magnetic fields typical of two assumed but representative electronic article surveillance (EAS) devices at 1 and 30 kHz, respectively. The devices assumed for the calculations are a solenoid type magnetic deactivator used at store checkouts and a pass-by panel-type EAS system consisting of two overlapping rectangular current-carrying coils used at entry and exit from a store. The impedance method code is modified to obtain induced current densities averaged over a cross section of 1 cm2 perpendicular to the direction of induced currents. This is done to compare the peak current densities with the limits or the basic restrictions given in the ICNIRP safety guidelines. Because of the stronger magnetic fields at lower heights for both the assumed devices, the peak 1 cm2 area-averaged current densities for the CNS tissues such as the brain and the spinal cord are increasingly larger for smaller models and are the highest for the model of the 5-year-old child. For both the EAS devices, the maximum 1 cm2 area-averaged current densities for the brain of the model of the adult are lower than the ICNIRP safety guideline, but may approach or exceed the ICNIRP basic restrictions for models of 10- and 5-year-old children if sufficiently strong magnetic fields are used.

  2. The transient electric field measurement system for EAST device

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Y., E-mail: wayong@ipp.ac.cn [Institute of Plasma Physics, Chinese Academy of Sciences, Hefei, Anhui (China); Ji, Z.S. [Institute of Plasma Physics, Chinese Academy of Sciences, Hefei, Anhui (China); Zhu, C.M. [The Experiment & Verification Center of State Grid Electric Power Research Institute (The Automation Equipment EMC Lab. of State Grid Co.), Nanjing, Jiangsu (China); Zhang, Z.C.; Ma, T.F.; Xu, Z.H. [Institute of Plasma Physics, Chinese Academy of Sciences, Hefei, Anhui (China)

    2016-11-15

    The electromagnetic environment around the Experimental Advanced Superconducting Tokamak (EAST) device is very complex during plasma discharge experiment. In order to fully monitor the changes of electric field around the EAST device during plasma discharge, a transient electric field measurement system based on PCI eXtensions for Instrumentation (PXI) platform has been designed. A digitizer is used for high-speed data acquisition of raw signals from electric field sensors, and a Field Programmable Gate Array (FPGA) module is used for realizing an on-the-fly fast Fourier transform (FFT) and inverse fast Fourier transform (IFFT) algorithm including a beforehand identified antenna factor (AF) to achieve finally a calibrated and filtered electric field measurement, then these signals can be displayed and easily analyzed. The raw signals from electric field sensors are transferred through optical fiber by optical isolation to reduce electromagnetic interference (EMI). The high speed data streaming technology is used for data storage. A prototype of this system has been realized to measure the transient electric field strength, with the real-time processing and continuous acquisition ability of one channel of 14-bit resolution and up to 50 MHz sampling rate, and 6 KHz FFT frequency resolution.

  3. Current density distribution mapping in PEM fuel cells as an instrument for operational measurements

    Energy Technology Data Exchange (ETDEWEB)

    Geske, M.; Heuer, M.; Heideck, G.; Styczynski, Z. A. [Otto-von-Guericke University Magdeburg, Chair Electric Power Networks and Renewable Energy Sources, Magdeburg (Germany)

    2010-07-01

    A newly developed measurement system for current density distribution mapping has enabled a new approach for operational measurements in proton exchange membrane fuel cells (PEMFC). Taking into account previously constructed measurement systems, a method based on a multi layer printed circuit board was chosen for the development of the new system. This type of system consists of a sensor, a special electronic device and the control and visualization PC. For the acquisition of the current density distribution values, a sensor device was designed and installed within a multilayer printed circuit board with integrated shunt resistors. Varying shunt values can be taken into consideration with a newly developed and evaluated calibration method. The sensor device was integrated in a PEM fuel cell stack to prove the functionality of the whole measurement system. A software application was implemented to visualize and save the measurement values. Its functionality was verified by operational measurements within a PEMFC system. Measurement accuracy and possible negative reactions of the sensor device during PEMFC operation are discussed in detail in this paper. The developed system enables operational measurements for different operating phases of PEM fuel cells. Additionally, this can be seen as a basis for new opportunities of optimization for fuel cell design and operation modes. (author)

  4. Current Density Distribution Mapping in PEM Fuel Cells as An Instrument for Operational Measurements

    Directory of Open Access Journals (Sweden)

    Martin Geske

    2010-04-01

    Full Text Available A newly developed measurement system for current density distribution mapping has enabled a new approach for operational measurements in proton exchange membrane fuel cells (PEMFC. Taking into account previously constructed measurement systems, a method based on a multi layer printed circuit board was chosen for the development of the new system. This type of system consists of a sensor, a special electronic device and the control and visualization PC. For the acquisition of the current density distribution values, a sensor device was designed and installed within a multilayer printed circuit board with integrated shunt resistors. Varying shunt values can be taken into consideration with a newly developed and evaluated calibration method. The sensor device was integrated in a PEM fuel cell stack to prove the functionality of the whole measurement system. A software application was implemented to visualize and save the measurement values. Its functionality was verified by operational measurements within a PEMFC system. Measurement accuracy and possible negative reactions of the sensor device during PEMFC operation are discussed in detail in this paper. The developed system enables operational measurements for different operating phases of PEM fuel cells. Additionally, this can be seen as a basis for new opportunities of optimization for fuel cell design and operation modes.

  5. Current-driven domain wall motion based memory devices: Application to a ratchet ferromagnetic strip

    Science.gov (United States)

    Sánchez-Tejerina, Luis; Martínez, Eduardo; Raposo, Víctor; Alejos, Óscar

    2018-04-01

    Ratchet memories, where perpendicular magnetocristalline anisotropy is tailored so as to precisely control the magnetic transitions, has been recently proven to be a feasible device to store and manipulate data bits. For such devices, it has been shown that the current-driven regime of domain walls can improve their performances with respect to the field-driven one. However, the relaxing time required by the traveling domain walls constitutes a certain drawback if the former regime is considered, since it results in longer device latencies. In order to speed up the bit shifting procedure, it is demonstrated here that the application of a current of inverse polarity during the DW relaxing time may reduce such latencies. The reverse current must be sufficiently high as to drive the DW to the equilibrium position faster than the anisotropy slope itself, but with an amplitude sufficiently low as to avoid DW backward shifting. Alternatively, it is possible to use such a reverse current to increase the proper range of operation for a given relaxing time, i.e., the pair of values of the current amplitude and pulse time that ensures single DW jumps for a certain latency time.

  6. SQUID based cryogenic current comparator for measurements of the dark current of superconducting cavities

    Energy Technology Data Exchange (ETDEWEB)

    Vodel, W.; Nietzsche, S.; Neubert, R.; Nawrodt, R. [Friedrich Schiller Univ. Jena (Germany); Peters, A. [GSI Darmstadt (Germany); Knaack, K.; Wendt, M.; Wittenburg, K. [DESY Hamburg (Germany)

    2005-07-01

    The linear accelerator technology, based on super-conducting L-band (1.3 GHz) is currently under study at DESY (Hamburg, Germany). The two 10 km long main Linacs will be equipped with a total of nearly 20.000 cavities. The dark current due to the emission of electrons in these high gradient field super-conducting cavities is an unwanted particle source. A newly high performance SQUID based measurement system for detecting dark currents is proposed. It makes use of the Cryogenic Current Comparator principle and senses dark currents in the pA range with a measurement bandwidth of up to 70 kHz. The use of a cryogenic current comparator as dark current sensor has some important advantages: -) the measurement of the absolute value of the dark current, -) the non-dependence on the electron trajectories, -) the accurate absolute calibration with an additional wire loop, and -) extremely high resolution.

  7. Influence of process parameters on threshold voltage and leakage current in 18nm NMOS device

    Science.gov (United States)

    Atan, Norani Binti; Ahmad, Ibrahim Bin; Majlis, Burhanuddin Bin Yeop; Fauzi, Izzati Binti Ahmad

    2015-04-01

    The process parameters are very crucial factor in the development of transistors. There are many process parameters that influenced in the development of the transistors. In this research, we investigate the effects of the process parameters variation on response characteristics such as threshold voltage (VTH) and sub-threshold leakage current (IOFF) in 18nm NMOS device. The technique to identify semiconductor process parameters whose variability would impact most on the device characteristic is realized through the process by using Taguchi robust design method. This paper presents the process parameters that influenced in threshold voltage (VTH) and sub-threshold leakage current (IOFF) which includes the Halo Implantation, Compensation Implantation, Adjustment Threshold voltage Implantation and Source/Drain Implantation. The design, fabrication and characterization of 18nm HfO2/TiSi2 NMOS device is simulated and performed via a tool called Virtual Wafer Fabrication (VWF) Silvaco TCAD Tool known as ATHENA and ATLAS simulators. These two simulators were combined with Taguchi L9 Orthogonal method to aid in the design and the optimization of the process parameters to achieve the optimum average of threshold voltage (VTH) and sub-threshold leakage current, (IOFF) in 18nm device. Results from this research were obtained; where Halo Implantation dose was identified as one of the process parameter that has the strongest effect on the response characteristics. Whereby the Compensation Implantation dose was identified as an adjustment factor to get the nominal values of threshold voltage VTH, and sub-threshold leakage current, IOFF for 18nm NMOS devices equal to 0.302849 volts and 1.9123×10-16 A/μm respectively. The design values are referred to ITRS 2011 prediction.

  8. Microfluidic device for cell capture and impedance measurement.

    Science.gov (United States)

    Jang, Ling-Sheng; Wang, Min-How

    2007-10-01

    This work presents a microfluidic device to capture physically single cells within microstructures inside a channel and to measure the impedance of a single HeLa cell (human cervical epithelioid carcinoma) using impedance spectroscopy. The device includes a glass substrate with electrodes and a PDMS channel with micro pillars. The commercial software CFD-ACE+ is used to study the flow of the microstructures in the channel. According to simulation results, the probability of cell capture by three micro pillars is about 10%. An equivalent circuit model of the device is established and fits closely to the experimental results. The circuit can be modeled electrically as cell impedance in parallel with dielectric capacitance and in series with a pair of electrode resistors. The system is operated at low frequency between 1 and 100 kHz. In this study, experiments show that the HeLa cell is successfully captured by the micro pillars and its impedance is measured by impedance spectroscopy. The magnitude of the HeLa cell impedance declines at all operation voltages with frequency because the HeLa cell is capacitive. Additionally, increasing the operation voltage reduces the magnitude of the HeLa cell because a strong electric field may promote the exchange of ions between the cytoplasm and the isotonic solution. Below an operating voltage of 0.9 V, the system impedance response is characteristic of a parallel circuit at under 30 kHz and of a series circuit at between 30 and 100 kHz. The phase of the HeLa cell impedance is characteristic of a series circuit when the operation voltage exceeds 0.8 V because the cell impedance becomes significant.

  9. Weak measurement from the electron displacement current: new path for applications

    International Nuclear Information System (INIS)

    Marian, D; Colomés, E; Oriols, X; Zanghì, N

    2015-01-01

    The interest on weak measurements is rapidly growing during the last years as a unique tool to better understand and predict new quantum phenomena. Up to now many theoretical and experimental weak-measurement techniques deal with (relativistic) photons or cold atoms, but there is much less investigation on (non-relativistic) electrons in up-to-date electronics technologies. We propose a way to perform weak measurements in nanoelectronic devices through the measurement of the total current (particle plus displacement component) in such devices. We study the interaction between an electron in the active region of a electron device with a metal surface working as a sensing electrode by means of the (Bohmian) conditional wave function. We perform numerical (Monte Carlo) simulations to reconstruct the Bohmian trajectories in the iconic double slit experiment. This work opens new paths for understanding the quantum properties of an electronic system as well as for exploring new quantum engineering applications in solid state physics. (paper)

  10. Device for measuring hole elongation in a bolted joint

    Science.gov (United States)

    Wichorek, Gregory R. (Inventor)

    1987-01-01

    A device to determine the operable failure mode of mechanically fastened lightweight composite joints by measuring the hole elongation of a bolted joint is disclosed. The double-lap joint test apparatus comprises a stud, a test specimen having a hole, two load transfer plates, and linear displacement measuring instruments. The test specimen is sandwiched between the two load transfer plates and clamped together with the stud. Spacer washers are placed between the test specimen and each load transfer plate to provide a known, controllable area for the determination of clamping forces around the hole of the specimen attributable to bolt torque. The spacer washers also provide a gap for the mounting of reference angles on each side of the test specimen. Under tensile loading, elongation of the hole of the test specimen causes the stud to move away from the reference angles. This displacement is measured by the voltage output of two linear displacement measuring instruments that are attached to the stud and remain in contact with the reference angles throughout the tensile loading. The present invention obviates previous problems in obtaining specimen deformation measurements by monitoring the reference angles to the test specimen and the linear displacement measuring instruments to the stud.

  11. Thin Film Energy Storage Device with Spray‐Coated Sliver Paste Current Collector

    Directory of Open Access Journals (Sweden)

    Seong Man Yoon

    2017-12-01

    Full Text Available This paper challenges the fabrication of a thin film energy storage device on a flexible polymer substrate specifically by replacing most commonly used metal foil current collectors with coated current collectors. Mass‐manufacturable spray‐coating technology enables the fabrication of two different half‐cell electric double layer capacitors (EDLC with a spray‐coated silver paste current collector and a Ni foil current collector. The larger specific capacitances of the half‐cell EDLC with the spray‐coated silver current collector are obtained as 103.86 F/g and 76.8 F/g for scan rates of 10 mV/s and 500 mV/s, respectively. Further, even though the half‐cell EDLC with the spray‐coated current collector is heavier than that with the Ni foil current collector, smaller Warburg impedance and contact resistance are characterized from Nyquist plots. For the applied voltages ranging from −0.5 V to 0.5 V, the spray‐coated thin film energy storage device exhibits a better performance.

  12. Design of a device for sky light polarization measurements.

    Science.gov (United States)

    Wang, Yujie; Hu, Xiaoping; Lian, Junxiang; Zhang, Lilian; Xian, Zhiwen; Ma, Tao

    2014-08-14

    Sky polarization patterns can be used both as indicators of atmospheric turbidity and as a sun compass for navigation. The objective of this study is to improve the precision of sky light polarization measurements by optimal design of the device used. The central part of the system is composed of a Charge Coupled Device (CCD) camera; a fish-eye lens and a linear polarizer. Algorithms for estimating parameters of the polarized light based on three images are derived and the optimal alignments of the polarizer are analyzed. The least-squares estimation is introduced for sky light polarization pattern measurement. The polarization patterns of sky light are obtained using the designed system and they follow almost the same patterns of the single-scattering Rayleigh model. Deviations of polarization angles between observation and the theory are analyzed. The largest deviations occur near the sun and anti-sun directions. Ninety percent of the deviations are less than 5° and 40% percent of them are less than 1°. The deviations decrease evidently as the degree of polarization increases. It also shows that the polarization pattern of the cloudy sky is almost identical as in the blue sky.

  13. Measuring device for water quality at reactor bottom

    Energy Technology Data Exchange (ETDEWEB)

    Urata, Hidehiro; Takagi, Jun-ichi

    1995-10-27

    The present invention concerns measurement for water quality at the bottom of a reactor of a BWR type plant, in which reactor water is sampled and analyzed in a state approximate to conditions in a pressure vessel. Based on the result, hydrogen injection amount is controlled during hydrogen injection operation. Namely, a monitor for water quality is disposed to a sampling line in communication with the bottom of a pressure vessel. A water quality monitor is disposed to a drain sampling line in communication with the bottom of the pressure vessel. A corrosion potentiometer is disposed to the pressure sampling line or the drain sampling line. A dissolved oxygen measuring device is disposed to the pressure vessel sampling line or the drain sampling line. With such a constitution, the reactor water can be sampled and analyzed in a state approximate to the conditions in the pressure vessel. In addition, signals from the water quality monitor are inputted to a hydrogen injection amount control device. As a result, the amount of hydrogen injected to primary coolants can be controlled in a state approximate to the conditions in the pressure vessel. (I.S.).

  14. Measuring device for water quality at reactor bottom

    International Nuclear Information System (INIS)

    Urata, Hidehiro; Takagi, Jun-ichi.

    1995-01-01

    The present invention concerns measurement for water quality at the bottom of a reactor of a BWR type plant, in which reactor water is sampled and analyzed in a state approximate to conditions in a pressure vessel. Based on the result, hydrogen injection amount is controlled during hydrogen injection operation. Namely, a monitor for water quality is disposed to a sampling line in communication with the bottom of a pressure vessel. A water quality monitor is disposed to a drain sampling line in communication with the bottom of the pressure vessel. A corrosion potentiometer is disposed to the pressure sampling line or the drain sampling line. A dissolved oxygen measuring device is disposed to the pressure vessel sampling line or the drain sampling line. With such a constitution, the reactor water can be sampled and analyzed in a state approximate to the conditions in the pressure vessel. In addition, signals from the water quality monitor are inputted to a hydrogen injection amount control device. As a result, the amount of hydrogen injected to primary coolants can be controlled in a state approximate to the conditions in the pressure vessel. (I.S.)

  15. Radioactivity concentration measuring device for radiation waste containing vessel

    International Nuclear Information System (INIS)

    Goto, Tetsuo.

    1994-01-01

    The device of the present invention can precisely and accurately measure a radioactive concentration of radioactive wastes irrespective of the radioactivity concentration distribution. Namely, a Ge detector having a collimator and a plurality of radiation detectors are placed at the outside of the radioactive waste containing vessel in such a way that it can rotate and move vertically relative to the vessel. The plurality of radiation detectors detect radiation coefficient signals at an assumed segment unit of a predetermined length in vertical direction and for every predetermined angle unit in the rotational direction. A weight measuring device determines the weight of the vessel. A computer calculates an average density of radioactivity for the region filled with radioactivity based on the determined net weight and radiation coefficient signals assuming that the volume of the radioactivity is constant. In addition, the computer calculates the amount of radioactivity in the assumed segment by conducting γ -ray absorption compensation calculation for the material in the vessel. Each of the amount of radioactivity is integrated to determine the amount of radioactivity in the vessel. (I.S.)

  16. Development of a safety case for the use of current limiting devices to manage short circuit currents on electrical distribution networks. Final report

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2004-07-01

    The original objective of this study was to review the safety issues associated with the use of current limiting devices and to write a risk assessment in accordance with good practice. But, when legislative procedures became apparent, the scope was changed to include involvement with the HSE, the DTI and Ofgem. It turned out that it would have been very difficult to write a safety case that would satisfy all of the agencies, or a risk assessment that would cover all applications. The scope of the study was therefore changed to focus on how the existing barriers should be tackled and the implications of the existing legislation. The approach to the study is described; it included reviews of background information and literature, questionnaires to manufacturers, a review of the reliability and hazards of the devices, and a review of UK safety legislation. The Final Report describes all this and includes discussion on the consequences of failure of fault current limiting devices, control measures which could be used to minimise risk, and recommendations for a way forward.

  17. Electron cyclotron beam measurement system in the Large Helical Device

    Energy Technology Data Exchange (ETDEWEB)

    Kamio, S., E-mail: kamio@nifs.ac.jp; Takahashi, H.; Kubo, S.; Shimozuma, T.; Yoshimura, Y.; Igami, H.; Ito, S.; Kobayashi, S.; Mizuno, Y.; Okada, K.; Osakabe, M.; Mutoh, T. [National Institute for Fusion Science, Toki 509-5292 (Japan)

    2014-11-15

    In order to evaluate the electron cyclotron (EC) heating power inside the Large Helical Device vacuum vessel and to investigate the physics of the interaction between the EC beam and the plasma, a direct measurement system for the EC beam transmitted through the plasma column was developed. The system consists of an EC beam target plate, which is made of isotropic graphite and faces against the EC beam through the plasma, and an IR camera for measuring the target plate temperature increase by the transmitted EC beam. This system is applicable to the high magnetic field (up to 2.75 T) and plasma density (up to 0.8 × 10{sup 19} m{sup −3}). This system successfully evaluated the transmitted EC beam profile and the refraction.

  18. Control and Data Analysis for Emittance Measuring Devices

    CERN Document Server

    Hoffmann, T

    2001-01-01

    Due to the wide range of heavy ion beam intensities and energies in the GSI linac and the associated transfer channel to the synchrotron, several different types of emittance measurement systems have been established. Many common devices such as slit/grid or dipole-sweep systems are integrated into the GSI control system. Other systems like the single shot pepper pot method using CCD-cameras or stand-alone slit/grid set-ups are connected to personal computers. An overview is given about the various systems and their software integration. Main interest is directed on the software development for emittance front-end control and data analysis such as evaluation algorithms or graphical presentation of the results. In addition, special features for improved usability of the software such as data export, project databases and automatic report generation will be presented. An outlook on a unified evaluation procedure for all different types of emittance measurement is given.

  19. Design and construction of a faraday cup for measuring small electron currents

    International Nuclear Information System (INIS)

    Veyssiere, A.

    1967-01-01

    This paper describes the design of a device for measuring and integrating very small currents generated by the impact of a charged particle beam upon a Faraday cup. Part one considers the detector as such. The main component is a graphite bloc capable of stopping practically all the incident charges. Part two describes the associated electrode apparatus required to measure better than 10 -13 ampere with a precision- of 1 per cent: Integration of such weak currents over periods of several hours, in the presence of a strong background current, is also discussed. (author) [fr

  20. Dose rate measuring device and dose rate measuring method using the same

    International Nuclear Information System (INIS)

    Urata, Megumu; Matsushita, Takashi; Hanazawa, Sadao; Konno, Takahiro; Chiba, Yoshinori; Yumitate, Tadahiro

    1998-01-01

    The device of the present invention comprises a scintillation fiber scope having a shape elongated in the direction of the height of a pressure vessel and emitting light by incident of radiation to detect radiation, a radioactivity measuring device for measuring a dose rate based on the detection of the fiber scope and a reel means for dispensing and taking up the fiber scope, and it constituted such that the dose rate of the pressure vessel and that of a shroud are determined independently. Then, when the taken out shroud is contained in an container, excessive shielding is not necessary, in addition, this device can reliably be inserted to or withdrawn from complicated places between the pressure vessel and the shroud, and further, the dose rate of the pressure vessel and that of the shroud can be measured approximately accurately even when the thickness of them is different greatly. (N.H.)

  1. Dose rate measuring device and dose rate measuring method using the same

    Energy Technology Data Exchange (ETDEWEB)

    Urata, Megumu; Matsushita, Takashi; Hanazawa, Sadao; Konno, Takahiro; Chiba, Yoshinori; Yumitate, Tadahiro

    1998-11-13

    The device of the present invention comprises a scintillation fiber scope having a shape elongated in the direction of the height of a pressure vessel and emitting light by incident of radiation to detect radiation, a radioactivity measuring device for measuring a dose rate based on the detection of the fiber scope and a reel means for dispensing and taking up the fiber scope, and it constituted such that the dose rate of the pressure vessel and that of a shroud are determined independently. Then, when the taken out shroud is contained in an container, excessive shielding is not necessary, in addition, this device can reliably be inserted to or withdrawn from complicated places between the pressure vessel and the shroud, and further, the dose rate of the pressure vessel and that of the shroud can be measured approximately accurately even when the thickness of them is different greatly. (N.H.)

  2. Electroluminescent efficiency of alternating current thick film devices using ZnS:Cu,Cl phosphor

    International Nuclear Information System (INIS)

    Sharma, Gaytri; Han, Sang Do; Kim, Jung Duk; Khatkar, Satyender P.; Rhee, Young Woo

    2006-01-01

    ZnS:Cu,Cl phosphor is prepared with the help of low intensity milling of the precursor material in two step firing process. The synthesized phosphor is used for the preparation of alternating current thick film electroluminescent (ACTFEL) devices with screen-printing method. The commission international de l'Eclairge (CIE) color co-ordinates of the ACTFEL devices prepared by these phosphor layers shows a shift from bluish-green to green region with the change in the amount of Cu in the phosphor. The various parameters to improve the efficiency and luminance of the devices have also been investigated. The brightness of the ac thick film EL device depends on the particle size of the phosphor, crystallinity, amount of binding material and applied voltage. The EL device fabricated with phosphor having average particle size of 25 μm shows maximum luminescence, when 60% phosphor concentration is used with respect to binding material. EL intensity is also linearly dependent on frequency. It is due the increase of excitation chances of the host matrix or dopant ions with increasing frequency

  3. Non-equilibrium properties of Josephson critical current in Nb-based three terminal superconducting tunnel devices

    International Nuclear Information System (INIS)

    Ammendola, G.; Parlato, L.; Peluso, G.; Pepe, G.

    1998-01-01

    Tunnel quasi-particle injection into a superconducting film provides useful information on the non-equilibrium state inside the perturbed superconductor as well as on the potential application to electronic devices. Three terminal injector-detector superconducting devices have a long history in non-equilibrium superconductivity. In the recent past non-equilibrium phenomena have attracted again considerable attention because of many superconducting based detectors involve processes substantially non-equilibrium in nature. The possibility of using a stacked double tunnel junction to study the influence of non-equilibrium superconductivity on the Josephson critical current is now considered. An experimental study of the effect of quasi-particle injection on the Josephson current both in steady-state and pulsed experiments down to T=1.2 K is presented using 3 terminal Nb-based stacked double tunnel devices. The feasibility of a new class of particle detectors based on the direct measurement of the change in the Josephson current following the absorption of a X-ray quantum is also discussed in terms of non-equilibrium theories. (orig.)

  4. Testing limits to airflow perturbation device (APD measurements

    Directory of Open Access Journals (Sweden)

    Jamshidi Shaya

    2008-10-01

    Full Text Available Abstract Background The Airflow Perturbation Device (APD is a lightweight, portable device that can be used to measure total respiratory resistance as well as inhalation and exhalation resistances. There is a need to determine limits to the accuracy of APD measurements for different conditions likely to occur: leaks around the mouthpiece, use of an oronasal mask, and the addition of resistance in the respiratory system. Also, there is a need for resistance measurements in patients who are ventilated. Method Ten subjects between the ages of 18 and 35 were tested for each station in the experiment. The first station involved testing the effects of leaks of known sizes on APD measurements. The second station tested the use of an oronasal mask used in conjunction with the APD during nose and mouth breathing. The third station tested the effects of two different resistances added in series with the APD mouthpiece. The fourth station tested the usage of a flexible ventilator tube in conjunction with the APD. Results All leaks reduced APD resistance measurement values. Leaks represented by two 3.2 mm diameter tubes reduced measured resistance by about 10% (4.2 cmH2O·sec/L for control and 3.9 cm H2O·sec/L for the leak. This was not statistically significant. Larger leaks given by 4.8 and 6.4 mm tubes reduced measurements significantly (3.4 and 3.0 cm cmH2O·sec/L, respectively. Mouth resistance measured with a cardboard mouthpiece gave an APD measurement of 4.2 cm H2O·sec/L and mouth resistance measured with an oronasal mask was 4.5 cm H2O·sec/L; the two were not significantly different. Nose resistance measured with the oronasal mask was 7.6 cm H2O·sec/L. Adding airflow resistances of 1.12 and 2.10 cm H2O·sec/L to the breathing circuit between the mouth and APD yielded respiratory resistance values higher than the control by 0.7 and 2.0 cm H2O·sec/L. Although breathing through a 52 cm length of flexible ventilator tubing reduced the APD

  5. The device for measuring amplitude spectra of ionizing irradiation

    International Nuclear Information System (INIS)

    Polyak, Yu.V.; Nebesnyj, A.F.

    1996-01-01

    The front-end electronic device for measuring of amplitude spectra of ionizing radiation have been made. The device have connection interface with the ionizing radiation detector, the pulse former, adapter, memory, electron ray tube with diode, supply unit and the regime setting unit of the work of electron ray tube. There are linear transmission scheme, level discriminator, pulse series - channel code converter, dividing capacitor in the device. Dynode of electron ray tube has been made in the form of rack or pads installed in the line parallel to axis of vertical scanning and electrically joined with each other. The distance between next tooth of rack or pads is Δy≥0,5 d, where d - diameter of focused electron beam of electron ray tube. The output of pulse former is joined with level discriminator and the first entrance of linear transmission scheme. The output of linear transmission scheme is joined with the entrance of pulse delay scheme, and the second entrance is joined with the first output of level discriminator. The output of pulse delay scheme is joined with Y-deflecting plate of electron ray tube. The first and the second entrance of pulse series - channel code converter are joined correspondingly with the output of adapter and the second output of level discriminator, and its output - with the entrance of memory unit. The first pin of loading resistor is joined through dividing capacitor with the entrance of adapter, its second pin - with the anode output nearest to dynode of electron ray tube. (E.V.Kh.)

  6. Indoor measurement of photovoltaic device characteristics at varying irradiance, temperature and spectrum for energy rating

    International Nuclear Information System (INIS)

    Bliss, M; Betts, T R; Gottschalg, R

    2010-01-01

    The first three-dimensional performance matrix for use in photovoltaic (PV) energy rating is reported utilizing a novel energy rating solar simulator based on LEDs. Device characteristics are measured indoors at varying irradiance (G), temperature (T) and spectrum (E). This opens the possibility for a more accurate measurement system for energy yield prediction of PV devices, especially for devices with high spectral dependence such as wide bandgap solar cells as they take into account spectral changes in the light. The main aspects of the LED-based solar simulator used are briefly described. A measurement method is developed and detailed in the paper, which takes into account the current imperfections in the achievable spectrum. Measurement results for a crystalline silicon solar cell are used to demonstrate the measurement approach. An uncertainty analysis of the measurement system is given, resulting in an overall absolute uncertainty of 4.3% (coverage factor k = 2) in maximum power measurements at 765 W m −2 irradiance with scope for further improvements

  7. An easy way to measure accurately the direct magnetoelectric voltage coefficient of thin film devices

    Energy Technology Data Exchange (ETDEWEB)

    Poullain, Gilles, E-mail: gilles.poullain@ensicaen.fr; More-Chevalier, Joris; Cibert, Christophe; Bouregba, Rachid

    2017-01-15

    Tb{sub x}Dy{sub 1−x}Fe{sub 2}/Pt/Pb(Zr{sub x}, Ti{sub 1−x})O{sub 3} thin films were grown on Pt/TiO{sub 2}/SiO{sub 2}/Si substrate by multi-target sputtering. The magnetoelectric voltage coefficient α{sup Η}{sub ΜΕ} was determined at room temperature using a lock-in amplifier. By adding, in series in the circuit, a capacitor of the same value as that of the device under test, we were able to demonstrate that the magnetoelectric device behaves as a voltage source. Furthermore, a simple way to subtract the stray voltage arising from the flow of eddy currents in the measurement set-up, is proposed. This allows the easy and accurate determination of the true magnetoelectric voltage coefficient. A large α{sup Η}{sub ΜΕ} of 8.3 V/cm. Oe was thus obtained for a Terfenol-D/Pt/PZT thin film device, without DC magnetic field nor mechanical resonance. - Highlights: • Magnetoelectric device behaves as a voltage source. • A simple way to subtract eddy currents during the measurement, is proposed.

  8. Stability of the superconductive operating mode in high current-density devices

    International Nuclear Information System (INIS)

    Wipf, S.L.

    1979-01-01

    The superconductive operating mode represents a thermal equilibrium that can tolerate a certain amount of disturbance before it is lost. The basin of attraction (BOA), in many ways equivalent to a potential well, is a measure of the size of disturbance needed to lift the device from the superconductive into a resistive operating mode. The BOA for a simple geometry is calculated and discussed. Experimental results are reported, showing how the concept is used to gain information on the disturbances occurring in a superconducting device

  9. Device for accurately measuring mass flow of gases

    Science.gov (United States)

    Hylton, James O.; Remenyik, Carl J.

    1994-01-01

    A device for measuring mass flow of gases which utilizes a substantially buoyant pressure vessel suspended within a fluid/liquid in an enclosure. The pressure vessel is connected to a weighing device for continuously determining weight change of the vessel as a function of the amount of gas within the pressure vessel. In the preferred embodiment, this pressure vessel is formed from inner and outer right circular cylindrical hulls, with a volume between the hulls being vented to the atmosphere external the enclosure. The fluid/liquid, normally in the form of water typically with an added detergent, is contained within an enclosure with the fluid/liquid being at a level such that the pressure vessel is suspended beneath this level but above a bottom of the enclosure. The buoyant pressure vessel can be interconnected with selected valves to an auxiliary pressure vessel so that initial flow can be established to or from the auxiliary pressure vessel prior to flow to or from the buoyant pressure vessel.

  10. Current measurement studies around the Cesme Peninsula (Turkey)

    International Nuclear Information System (INIS)

    Taspinar, N.

    1989-04-01

    In order to design coastal structures and marine vehicles safely, it is required to know current climate which shows the variation of the current characteristics with time. There are a wide variety of current meters designed to measure water flow today. Each current meter is capable of recording the influence of mooring arrangement. Here we describe sea water temperatures, salinities and current velocities at offshore of Akburun, Tatlicak Burnu, Kalem Burnu and Kizil Burun areas in Cesme Peninsula 27 August, 1986 to 19 November, 1986. At the end of the investigations, measured significant maximum and average current velocities have been routinely analysed with micro-computers and also the percentages of current velocity have been calculated. (author). 8 refs, 6 figs, 4 tabs

  11. On the accuracy of current TCAD hot carrier injection models in nanoscale devices

    Science.gov (United States)

    Zaka, Alban; Rafhay, Quentin; Iellina, Matteo; Palestri, Pierpaolo; Clerc, Raphaël; Rideau, Denis; Garetto, Davide; Dornel, Erwan; Singer, Julien; Pananakakis, Georges; Tavernier, Clément; Jaouen, Hervé

    2010-12-01

    In this work, the hot electron injection models presently available for technology support have been investigated within the context of the development of advanced embedded non-volatile memories. The distribution functions obtained by these models (namely the Fiegna Model - FM [1], the Lucky Electron Model - LEM [2] and the recently implemented Spherical Harmonics Expansion of the Boltzman's Transport Equation - SHE [3]), have been systematically compared to rigorous Monte Carlo (MC) results [4], both in homogeneous and device conditions. Gate-to-drain current ratio and gate current density simulation has also been benchmarked in device simulations. Results indicate that local models such as FM, can partially capture the channel hot electron injection, at the price of model parameter adjustments. Moreover, at least in the device and field condition considered in this work, an overall better agreement with MC simulations has been obtained using the 1st order SHE, even without any particular fitting procedure. Extending the results presented in [3] by exploring shorter gate lengths and addressing the floating gate voltage dependence of the gate current, this work shows that the SHE method could contribute to bridge the gap between the rigorous but time consuming MC method and less rigorous but suitable TCAD local models.

  12. Study on ICT specification devices compared with needs and current technologies at Nuclear Malaysia

    International Nuclear Information System (INIS)

    Mohd Fauzi Haris; Raja Murzaferi Raja Moktar; Mohd Hafez Mohd Tahir

    2012-01-01

    In line with current development of ICT, Malaysian government has planned and introduced several initiatives based on ICT strategies. In Economic Transformation Programs, these matters were mentioned in Chapter 13 entitled Communication Content and Infrastructure. In order to make these plans successful, sustainability and preparedness of ICT are required. ICT devices were not focused only on computer but also others components that supported and increased the performance of computer itself. This paper discussed on data produced from study of current ICT needs in line with technology and in future hopefully it can support all the planning made by the government. (author)

  13. A Novel Electronic Device for Measuring Urine Flow Rate: A Clinical Investigation

    Directory of Open Access Journals (Sweden)

    Aliza Goldman

    2017-09-01

    Full Text Available Objective: Currently, most vital signs in the intensive care unit (ICU are electronically monitored. However, clinical practice for urine output (UO measurement, an important vital sign, usually requires manual recording of data that is subject to human errors. In this study, we assessed the ability of a novel electronic UO monitoring device to measure real-time hourly UO versus current clinical practice. Design: Patients were connected to the RenalSense Clarity RMS Sensor Kit with a sensor integrated within a standard sterile urinary catheter drainage tube to monitor urine flow in real time. The Clarity RMS Sensor Kit was modified to incorporate a standard urinometer (Unomedical for the nursing staff to record UO as per their standard practice. The drainage bag was placed in a container on a scientific scale (Precisa BJ to be used as the gold standard. Interventions: Nursing records for hourly UO were collected and compared with the electronically recorded UO. Sensor measurements and nursing staff manual records of UO were compared with the scale data. Setting: The study setting was the ICU of Hadassah Hospital, Jerusalem. Patients: Data from 23 patients with a urinary catheter were observed in this study. Measurements and main results: A total of 1046 hours of UO were recorded from 23 subjects. Compared with the scale data, the measurements of hourly urine flow measured with the RenalSense system were closer, had a better correlation, and narrower limits of agreement to gravimetrically determined values than the measurements obtained by the nurses. In addition, continuous monitoring of UO provided graphical display of response to repeated diuretic administration. Conclusions: An electronic device for recording UO has been shown to provide more reliable information of UO records and patient fluid status than current practice. Future applications of this device will provide valuable information to help set protocol goals such as decisions for

  14. One-sided measurement-device-independent quantum key distribution

    Science.gov (United States)

    Cao, Wen-Fei; Zhen, Yi-Zheng; Zheng, Yu-Lin; Li, Li; Chen, Zeng-Bing; Liu, Nai-Le; Chen, Kai

    2018-01-01

    Measurement-device-independent quantum key distribution (MDI-QKD) protocol was proposed to remove all the detector side channel attacks, while its security relies on the trusted encoding systems. Here we propose a one-sided MDI-QKD (1SMDI-QKD) protocol, which enjoys detection loophole-free advantage, and at the same time weakens the state preparation assumption in MDI-QKD. The 1SMDI-QKD can be regarded as a modified MDI-QKD, in which Bob's encoding system is trusted, while Alice's is uncharacterized. For the practical implementation, we also provide a scheme by utilizing coherent light source with an analytical two decoy state estimation method. Simulation with realistic experimental parameters shows that the protocol has a promising performance, and thus can be applied to practical QKD applications.

  15. Low-temperature mobility measurements on CMOS devices

    International Nuclear Information System (INIS)

    Hairpetian, A.; Gitlin, D.; Viswanathan, C.R.

    1989-01-01

    The surface channel mobility of carriers in eta- and rho-MOS transistors fabricated in a CMOS process was accurately determined at low temperatures down to 5 Κ. The mobility was obtained by an accurate measurement of the inversion charge density using a split C-V technique and the conductance at low drain voltages. The split C-V technique was validated at all temperatures using a one-dimensional Poisson solver (MOSCAP), which was modified for low-temperature application. The mobility dependence on the perpendicular electric field for different substrate bias values appears to have different temperature dependence for eta- and rho-channel devices. The electron mobility increases with a decrease in temperature at all gate voltages. On the other hand, the hole mobility exhibits a different temperature behavior depending upon whether the gate voltage corresponds to strong inversion or is near threshold

  16. A Method for Eddy Current Field Measurement in Permanent Magnet Magnetic Resonance Imaging Systems

    Directory of Open Access Journals (Sweden)

    SONG Rui

    2018-03-01

    Full Text Available Magnetic resonance imaging (MRI is a widely used medical imaging technique. In MRI system, gradient magnetic fields are used to code spatial information. However, the fast-switching electric currents in the gradients coils used to generate gradient fields also induce vortex electric field, often referred as eddy current, in the surrounding metal conductors. In this paper, a method for eddy current field measurement was proposed. Based on the Faraday law of electromagnetic induction, an eddy current field measuring device was designed. Combining hardware acquisition and software processing, the eddy current field was obtained by subtracting the ideal gradient field from the magnetic field measured experimentally, whose waveform could be displayed in real time. The proposed method was verified by experimental results.

  17. Basic properties of the current-current correlation measure for random Schroedinger operators

    International Nuclear Information System (INIS)

    Hislop, Peter D.; Lenoble, Olivier

    2006-01-01

    The current-current correlation measure plays a crucial role in the theory of conductivity for disordered systems. We prove a Pastur-Shubin-type formula for the current-current correlation measure expressing it as a thermodynamic limit for random Schroedinger operators on the lattice and the continuum. We prove that the limit is independent of the self-adjoint boundary conditions and independent of a large family of expanding regions. We relate this finite-volume definition to the definition obtained by using the infinite-volume operators and the trace-per-unit volume

  18. Clinical assessment of the accuracy of blood glucose measurement devices.

    Science.gov (United States)

    Pfützner, Andreas; Mitri, Michael; Musholt, Petra B; Sachsenheimer, Daniela; Borchert, Marcus; Yap, Andrew; Forst, Thomas

    2012-04-01

    Blood glucose meters for patient self-measurement need to comply with the accuracy standards of the ISO 15197 guideline. We investigated the accuracy of the two new blood glucose meters BG*Star and iBG*Star (Sanofi-Aventis) in comparison to four other competitive devices (Accu-Chek Aviva, Roche Diagnostics; FreeStyle Freedom Lite, Abbott Medisense; Contour, Bayer; OneTouch Ultra 2, Lifescan) at different blood glucose ranges in a clinical setting with healthy subjects and patients with type 1 and type 2 diabetes. BGStar and iBGStar are employ dynamic electrochemistry, which is supposed to result in highly accurate results. The study was performed on 106 participants (53 female, 53 male, age (mean ± SD): 46 ± 16 years, type 1: 32 patients, type 2: 34 patients, and 40 healthy subjects). Two devices from each type and strips from two different production lots were used for glucose assessment (∼200 readings/meter). Spontaneous glucose assessments and glucose or insulin interventions under medical supervision were applied to perform measurements in the different glucose ranges in accordance with the ISO 15197 requirements. Sample values 400 mg/dL were prepared by laboratory manipulations. The YSI glucose analyzer (glucose oxidase method) served as the standard reference method which may be considered to be a limitation in light of glucose hexokinase-based meters. For all devices, there was a very close correlation between the glucose results compared to the YSI reference method results. The correlation coefficients were r = 0.995 for BGStar and r = 0.992 for iBGStar (Aviva: 0.995, Freedom Lite: 0.990, Contour: 0.993, Ultra 2: 0.990). Error-grid analysis according to Parkes and Clarke revealed both 100% of the readings to be within the clinically acceptable areas (Clarke: A + B with BG*Star (100 + 0), Aviva (97 + 3), and Contour (97 + 3); and 99.5% with iBG*Star (97.5 + 2), Freedom Lite (98 + 1.5), and Ultra 2 (97.5 + 2

  19. Mobile Communication Devices, Ambient Noise, and Acoustic Voice Measures.

    Science.gov (United States)

    Maryn, Youri; Ysenbaert, Femke; Zarowski, Andrzej; Vanspauwen, Robby

    2017-03-01

    The ability to move with mobile communication devices (MCDs; ie, smartphones and tablet computers) may induce differences in microphone-to-mouth positioning and use in noise-packed environments, and thus influence reliability of acoustic voice measurements. This study investigated differences in various acoustic voice measures between six recording equipments in backgrounds with low and increasing noise levels. One chain of continuous speech and sustained vowel from 50 subjects with voice disorders (all separated by silence intervals) was radiated and re-recorded in an anechoic chamber with five MCDs and one high-quality recording system. These recordings were acquired in one condition without ambient noise and in four conditions with increased ambient noise. A total of 10 acoustic voice markers were obtained in the program Praat. Differences between MCDs and noise condition were assessed with Friedman repeated-measures test and posthoc Wilcoxon signed-rank tests, both for related samples, after Bonferroni correction. (1) Except median fundamental frequency and seven nonsignificant differences, MCD samples have significantly higher acoustic markers than clinical reference samples in minimal environmental noise. (2) Except median fundamental frequency, jitter local, and jitter rap, all acoustic measures on samples recorded with the reference system experienced significant influence from room noise levels. Fundamental frequency is resistant to recording system, environmental noise, and their combination. All other measures, however, were impacted by both recording system and noise condition, and especially by their combination, often already in the reference/baseline condition without added ambient noise. Caution is therefore warranted regarding implementation of MCDs as clinical recording tools, particularly when applied for treatment outcomes assessments. Copyright © 2017 The Voice Foundation. Published by Elsevier Inc. All rights reserved.

  20. Low noise constant current source for bias dependent noise measurements

    International Nuclear Information System (INIS)

    Talukdar, D.; Bose, Suvendu; Bardhan, K. K.; Chakraborty, R. K.

    2011-01-01

    A low noise constant current source used for measuring the 1/f noise in disordered systems in ohmic as well as nonohmic regime is described. The source can supply low noise constant current starting from as low as 1 μA to a few tens of milliampere with a high voltage compliance limit of around 20 V. The constant current source has several stages, which can work in a standalone manner or together to supply the desired value of load current. The noise contributed by the current source is very low in the entire current range. The fabrication of a low noise voltage preamplifier modified for bias dependent noise measurements and based on the existing design available in the MAT04 data sheet is also described.

  1. Measurement of Neutrino Induced, Charged Current, Charged Pion Production

    Energy Technology Data Exchange (ETDEWEB)

    Wilking, Michael Joseph [Univ. of Colorado, Boulder, CO (United States)

    2009-05-01

    Neutrinos are among the least understood particles in the standard model of particle physics. At neutrino energies in the 1 GeV range, neutrino properties are typically determined by observing the outgoing charged lepton produced in a charged current quasi-elastic interactions. The largest charged current background to these measurements comes from charged current pion production interactions, for which there is very little available data.

  2. Hall probe for measuring high currents in superconducting coils

    International Nuclear Information System (INIS)

    Ferendeci, A.M.

    1986-01-01

    Constructional details of a compact Hall probe for measuring high currents in superconducting coils are given. The Hall probe is easy to assemble and can be inserted or removed from the system without breaking the superconducting loop. Upper current limit of the probe can be increased by using larger magnetic core material. Shielding becomes necessary if the probe holder is to be placed near large current dependent magnetic fields

  3. 16 CFR Figures 1 and 2 to Part 1204 - Suggested Instrumentation for Current Monitoring Device and High Voltage Facility

    Science.gov (United States)

    2010-01-01

    ... 16 Commercial Practices 2 2010-01-01 2010-01-01 false Suggested Instrumentation for Current Monitoring Device and High Voltage Facility 1 Figures 1 and 2 to Part 1204 Commercial Practices CONSUMER... Instrumentation for Current Monitoring Device and High Voltage Facility EC03OC91.008 ...

  4. Converter for Measurement of non-sinusoidal current peak value

    DEFF Research Database (Denmark)

    Butvin, P.; Nielsen, Otto V; Brauer, Peter

    1997-01-01

    A linear-response toroid with core wound of rapidly quenched soft magnetic metallic ribbon and fitted with two windings is used to enable correct measurement of mean peak value of non-sinusoidal and not noise-free alternating current.......A linear-response toroid with core wound of rapidly quenched soft magnetic metallic ribbon and fitted with two windings is used to enable correct measurement of mean peak value of non-sinusoidal and not noise-free alternating current....

  5. Dark Current And Voltage Measurements Of Metal-Organic-Semiconductor (M-Or-S) Diode

    International Nuclear Information System (INIS)

    Adianto

    1996-01-01

    . Some Metal-Organic-Semiconductor (M-Or-S) thin film diodes, constructed with an organic polymer (polymerized toluene) as an active component has been successfully fabricated. The thin film M-Or-S diodes were fabricated on an n-type silicon with resistivity of 250-500 Ocm and p type silicon with resistivity of 10-20 Ocm as a substrate with polymerized toluene used as insulator. When deposited on silicon wafers with electrode of evaporated Ni on the n-type silicon and evaporated Au as the electrode on the polymerized toluene film, the electronic devices of Metal-Organic- Semiconductor (M-Or-S) type can be produced with one of its characteristics is that their light sensitivity. A plasma ion deposition system was constructed and used to deposit organic monomeric substance (toluene) that functioned as an isolator between semiconductor and the evaporated metal electrodes. The current-voltage measurements for different configurations of M-Or-S devices were carried out to determine the current-voltage (1-V) characteristics for M-Or-S devices with different materials and thicknesses. In addition to the 1-V measurement mentioned before, 1-V measurements of the devices were also carried out by using a curve tracer oscilloscope, and the picture of the effective parameters of each of the device could be taken by using a polaroid camera. Since the devices are very sensitive to light, the devices were all tested in a black-box which was covered by a black cloth to make sure that there was no light coming through. The experimental results for p- and n-type silicon substrates showed that an M-Or-S diode with n-type gave a higher breakdown voltage than that p- type silicon. In addition, the reverse bias breakdown voltage increased as the thickness of the thin film increased in the range of 50 -2500 V/μm

  6. A Systematic Review on Existing Measures for the Subjective Assessment of Rehabilitation and Assistive Robot Devices

    Directory of Open Access Journals (Sweden)

    Yiannis Koumpouros

    2016-01-01

    Full Text Available The objective of the current study is to identify and classify outcome measures currently used for the assessment of rehabilitation or assistive robot devices. We conducted a systematic review of the literature using PubMed, MEDLINE, CIRRIE, and Scopus databases for studies that assessed rehabilitation or assistive robot devices from 1980 through January 2016. In all, 31 articles met all inclusion criteria. Tailor-made questionnaires were the most commonly used tool at 66.7%, while the great majority (93.9% of the studies used nonvalidated instruments. The study reveals the absence of a standard scale which makes it difficult to compare the results from different researchers. There is a great need, therefore, for a valid and reliable instrument to be available for use by the intended end users for the subjective assessment of robot devices. The study concludes by identifying two scales that have been validated in general assistive technology devices and could support the scope of subjective assessment in rehabilitation or assistive robots (however, with limited coverage and a new one called PYTHEIA, recently published. The latter intends to close the gap and help researchers and developers to evaluate, assess, and produce products that satisfy the real needs of the end users.

  7. A Fiber-Optic Aircraft Lightning Current Measurement Sensor

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George N.

    2013-01-01

    A fiber-optic current sensor based on the Faraday Effect is developed for aircraft installations. It can measure total lightning current amplitudes and waveforms, including continuing current. Additional benefits include being small, lightweight, non-conducting, safe from electromagnetic interference, and free of hysteresis and saturation. The Faraday Effect causes light polarization to rotate in presence of magnetic field in the direction of light propagation. Measuring the total induced light polarization change yields the total current enclosed. The system operates at 1310nm laser wavelength and can measure approximately 300 A - 300 kA, a 60 dB range. A reflective polarimetric scheme is used, where the light polarization change is measured after a round-trip propagation through the fiber. A two-detector setup measures the two orthogonal polarizations for noise subtraction and improved dynamic range. The current response curve is non-linear and requires a simple spline-fit correction. Effects of high current were achieved in laboratory using combinations of multiple fiber and wire loops. Good result comparisons against reference sensors were achieved up to 300 kA. Accurate measurements on a simulated aircraft fuselage and an internal structure illustrate capabilities that maybe difficult with traditional sensors. Also tested at a commercial lightning test facility from 20 kA to 200 kA, accuracy within 3-10% was achieved even with non-optimum setups.

  8. Measurements of current penetration during PDX discharge start-up

    International Nuclear Information System (INIS)

    Meyerhofer, D.D.; Goldston, R.J.; Kaita, R.; Cavallo, A.; Grek, B.; Johnson, D.; McCune, D.C.; McGuire, K.; White, R.B.

    1984-11-01

    The current penetration phase of PDX discharges is examined. The Fast Ion Diagnostic Experiment has been used to measure the temporal evolution of the central q (r/a < 0.4), and to show the effect of magnetic perturbations on fast ions. During plasma current penetration, a series of magnetic perturbations was observed in the plasma. If the current was rising rapidly, the perturbations were accompanied by increases in β/sub theta/ + l/sub i//2 and decreases in the loop voltage, suggesting a rapid penetration of the plasma current. When the plasma current was rising slowly, a series of minor disruptions occurred. These were accompanied by decreases in β/sub theta/ + l/sub i//2 and the loop voltage, and increases in the plasma current. During this phase, current penetration may be enhanced by the change in the resistivity profile which accompanies the disruption

  9. Test measurements on the RF charge breeder device BRIC

    International Nuclear Information System (INIS)

    Variale, Vincenzo; Boggia, Antonio; Clauser, Tarcisio; Raino, Antonio; Valentino, Vincenzo; Verrone, Grazia; Bak, Petr; Kustenzov, Gennady; Skarbo, Boris; Tiunov, Michael

    2004-01-01

    The 'charge state breeder' BRIC (BReeding Ion Charge) is based on an EBIS source and it is designed to accept Radioactive Ion Beam (RIB) with charge state +1, in a slow injection mode, to increase their charge state up to +n. BRIC has been developed at the INFN section of Bari (Italy) during these last 3 years with very limited funds. Now, it has been assembled at the LNL (Italy) where are in progress the first tests as stand alone source and where, in the future, with some implementation, it will be tested as charge breeder at ISOL/TS facility of that laboratory. BRIC could be considered as a solution for the charge state breeder of the SPES project under study also at the LNL. The new feature of BRIC, with respect to the classical EBIS, is given by the insertion, in the ion drift chamber, of a radio frequency (RF) - quadrupole aiming to filter the unwanted masses and then making a more efficient containment of the wanted ions. In this paper, the first ion charge state measurements and analysis and the effect of the RF field applied on the ion chamber will be reported and discussed. The first RF test measurements seem confirm, as foreseen by simulation results carried out previously, that a selective containment can be obtained. However, most accurate measurements needed to study with more details the effect. For this reason, few implementations of the system are in order to improve the accuracy of the measurements. The proposed modifications of the BRIC device, then, will be also presented and shortly discussed

  10. Measurements of the Exerted Pressure by Pelvic Circumferential Compression Devices

    Science.gov (United States)

    Knops, Simon P; van Riel, Marcel P.J.M; Goossens, Richard H.M; van Lieshout, Esther M.M; Patka, Peter; Schipper, Inger B

    2010-01-01

    Background: Data on the efficacy and safety of non-invasive Pelvic Circumferential Compression Devices (PCCDs) is limited. Tissue damage may occur if a continuous pressure on the skin exceeding 9.3 kPa is sustained for more than two or three hours. The aim of this study was to gain insight into the pressure build-up at the interface, by measuring the PCCD-induced pressure when applying pulling forces to three different PCCDs (Pelvic Binder® , SAM-Sling ® and T-POD® ) in a simplified model. Methods: The resulting exerted pressures were measured at four ‘anatomical’ locations (right, left, posterior and anterior) in a model using a pressure measurement system consisting of pressure cuffs. Results: The exerted pressure varied substantially between the locations as well as between the PCCDs. Maximum pressures ranged from 18.9-23.3 kPa and from 19.2-27.5 kPa at the right location and left location, respectively. Pressures at the posterior location stayed below 18 kPa. At the anterior location pressures varied markedly between the different PCCDs. Conclusion: The circumferential compression by the different PCCDs showed high pressures measured at the four locations using a simplified model. Difference in design and functional characteristics of the PCCDs resulted in different pressure build-up at the four locations. When following the manufacturer’s instructions, the exerted pressure of all three PCCDs tested exceeded the tissue damaging level (9.3 kPa). In case of prolonged use in a clinical situation this might put patients at risk for developing tissue damage. PMID:20361001

  11. Evaluation of spectrum measurement devices for operational use

    CERN Document Server

    Devine, R T; Gray, D W; Seagraves, D T; Olsher, R H; Johnson, J P

    2002-01-01

    Several neutron spectrometers manufactured by Bubble Technology Industries (BTI) were tested and evaluated in a variety of neutron fields. Findings and conclusions are presented for the following BTI instruments: a modification of the Rotational Spectrometer (ROSPEC) that includes a thermal and epithermal capability, the Simple Scintillation Spectrometer that is used in conjunction with the ROSPEC to extend its high-energy range, and the MICROSPEC N-Probe which is capable of providing a crude spectrum over the energy range from thermal to 18 MeV. The main objective of these measurements was to determine the accuracy of both the energy spectrum and dose equivalent information generated by these devices. In addition, the dose response of the Wide-Energy Neutron Detection Instrument (WENDI-II) was measured in all neutron fields relative to a bare sup 2 sup 5 sup 2 Cf calibration. The performance of the WENDI-II rem meter was compared to the dose information generated by the neutron spectrometers. The instruments...

  12. Are wearable devices ready for HTTPS? Measuring the cost of secure communication protocols on wearable devices

    OpenAIRE

    Kolamunna, Harini; Chauhan, Jagmohan; Hu, Yining; Thilakarathna, Kanchana; Perino, Diego; Makaroff, Dwight; Seneviratne, Aruna

    2016-01-01

    The majority of available wearable devices require communication with Internet servers for data analysis and storage, and rely on a paired smartphone to enable secure communication. However, wearable devices are mostly equipped with WiFi network interfaces, enabling direct communication with the Internet. Secure communication protocols should then run on these wearables itself, yet it is not clear if they can be efficiently supported. In this paper, we show that wearable devices are ready for...

  13. Graphene-Nanodiamond Heterostructures and their application to High Current Devices

    Science.gov (United States)

    Zhao, Fang; Vrajitoarea, Andrei; Jiang, Qi; Han, Xiaoyu; Chaudhary, Aysha; Welch, Joseph O.; Jackman, Richard B.

    2015-01-01

    Graphene on hydrogen terminated monolayer nanodiamond heterostructures provides a new way to improve carrier transport characteristics of the graphene, offering up to 60% improvement when compared with similar graphene on SiO2/Si substrates. These heterostructures offers excellent current-carrying abilities whilst offering the prospect of a fast, low cost and easy methodology for device applications. The use of ND monolayers is also a compatible technology for the support of large area graphene films. The nature of the C-H bonds between graphene and H-terminated NDs strongly influences the electronic character of the heterostructure, creating effective charge redistribution within the system. Field effect transistors (FETs) have been fabricated based on this novel herterostructure to demonstrate device characteristics and the potential of this approach. PMID:26350107

  14. Python Scripts for Automation of Current-Voltage Testing of Semiconductor Devices (FY17)

    Science.gov (United States)

    2017-01-01

    investment required by the user with manual operation, only a fraction of the total devices fabricated on a sample are actually tested. In this...only moves the stage itself in 3 dimensions. It does not control the probes or stage rotation. It is important to note that the user must first attempt...measurements using the 4155C and the manual probe station. This script bypasses the original front panel operation of the 4155C and allows the user to set

  15. [Mobile Health: IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices].

    Science.gov (United States)

    Zhou, Xia; Wu, Wenli; Bao, Shudi

    2015-07-01

    IEEE Std 1708-2014 breaks through the traditional standards of cuff based blood pressure measuring devices and establishes a normative definition of wearable cuffless blood pressure measuring devices and the objective performance evaluation of this kind of devices. This study firstly introduces the background of the new standard. Then, the standard details will be described, and the impact of cuffless blood pressure measuring devices with the new standard on manufacturers and end users will be addressed.

  16. Eddy-current flow rate meter for measuring sodium flow rates

    International Nuclear Information System (INIS)

    Knaak, J.

    1976-01-01

    For safety reasons flow rate meters for monitoring coolant flow rates are inserted in the core of sodium-cooled fast breeder reactors. These are so-called eddy-current flow rate meters which can be mounted directly above the fuel elements. In the present contribution the principle of measurement, the mechanical construction and the circuit design of the flow rate measuring device are described. Special problems and their solution on developing the measuring system are pointed out. Finally, results of measurement and experience with the apparatus in several experiments are reported, where also further possibilities of application were tested. (orig./TK) [de

  17. Fiber-Optic Sensor for Aircraft Lightning Current Measurement

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George G.; Mata, Carlos T.; Mata, Angel G.; Snyder, Gary P.

    2012-01-01

    An electric current sensor based on Faraday rotation effect in optical fiber was developed for measuring aircraft lightning current. Compared to traditional sensors, the design has many advantages including the ability to measure total current and to conform to structure geometries. The sensor is also small, light weight, non-conducting, safe from interference, and free of hysteresis and saturation. Potential applications include characterization of lightning current waveforms, parameters and paths, and providing environmental data for aircraft certifications. In an optical fiber as the sensing medium, light polarization rotates when exposed to a magnetic field in the direction of light propagation. By forming closed fiber loops around a conductor and applying Ampere s law, measuring the total light rotation yields the enclosed current. A reflective polarimetric scheme is used, where polarization change is measured after the polarized light travels round-trip through the sensing fiber. The sensor system was evaluated measuring rocket-triggered lightning over the 2011 summer. Early results compared very well against a reference current shunt resistor, demonstrating the sensor s accuracy and feasibility in a lightning environment. While later comparisons show gradually increasing amplitude deviations for an undetermined cause, the overall waveforms still compared very well.

  18. A silicon-on-insulator vertical nanogap device for electrical transport measurements in aqueous electrolyte solution

    Energy Technology Data Exchange (ETDEWEB)

    Strobel, Sebastian [Walter Schottky Institut, Technische Universitaet Muenchen, Am Coulombwall, D-85748 Garching (Germany); Arinaga, Kenji [Walter Schottky Institut, Technische Universitaet Muenchen, Am Coulombwall, D-85748 Garching (Germany); Hansen, Allan [Walter Schottky Institut, Technische Universitaet Muenchen, Am Coulombwall, D-85748 Garching (Germany); Tornow, Marc [Walter Schottky Institut, Technische Universitaet Muenchen, Am Coulombwall, D-85748 Garching (Germany)

    2007-07-25

    A novel concept for metal electrodes with few 10 nm separation for electrical conductance measurements in an aqueous electrolyte environment is presented. Silicon-on-insulator (SOI) material with 10 nm buried silicon dioxide serves as a base substrate for the formation of SOI plateau structures which, after recess-etching the thin oxide layer, thermal oxidation and subsequent metal thin film evaporation, feature vertically oriented nanogap electrodes at their exposed sidewalls. During fabrication only standard silicon process technology without any high-resolution nanolithographic techniques is employed. The vertical concept allows an array-like parallel processing of many individual devices on the same substrate chip. As analysed by cross-sectional TEM analysis the devices exhibit a well-defined material layer architecture, determined by the chosen material thicknesses and process parameters. To investigate the device in aqueous solution, we passivated the sample surface by a polymer layer, leaving a micrometre-size fluid access window to the nanogap region only. First current-voltage characteristics of a 65 nm gap device measured in 60 mM buffer solution reveal excellent electrical isolation behaviour which suggests applications in the field of biomolecular electronics in a natural environment.

  19. Online junction temperature measurement using peak gate current

    DEFF Research Database (Denmark)

    Baker, Nick; Munk-Nielsen, Stig; Iannuzzo, Francesco

    2015-01-01

    A new method for junction temperature measurement of MOS-gated power semiconductor switches is presented. The measurement method involves detecting the peak voltage over the external gate resistor of an IGBT or MOSFET during turn-on. This voltage is directly proportional to the peak gate current...

  20. Bigger, Brighter, Bluer-Better?Current light-emitting devices- adverse sleep properties and preventative strategies.

    Directory of Open Access Journals (Sweden)

    Paul eGringras

    2015-10-01

    Full Text Available ObjectiveIn an effort to enhance the efficiency, brightness and contrast of light-emitting (LE devices during the day, displays often generate substantial short-wavelength (blue-enriched light emissions that can adversely affect sleep. We set out to verify the extent of such short-wavelength emissions, produced by a tablet (iPad Air, e-reader (Kindle Paperwhite 1st generation and smartphone (iPhone 5s and to determine the impact of strategies designed to reduce these light emissions. SettingUniversity of Surrey dedicated chronobiology facility.MethodsFirstly, the spectral power of all the light-emitting (LE devices was assessed when displaying identical text. Secondly, we compared the text output with that of ‘Angry Birds’-a popular top 100 ‘App Store’ game. Finally we measured the impact of two strategies that attempt to reduce the output of short-wavelength light emissions. The first strategy employed an inexpensive commercially available pair of orange-tinted ‘blue-blocking’ glasses. The second tested an app designed to be ‘sleep-aware’ whose designers deliberately attempted to reduce blue-enriched light emissions.ResultsAll the LE devices shared very similar enhanced blue-light peaks when displaying text. This included the output from the backlit Kindle Paperwhite device. The spectra when comparing text to the Angry Birds game were also very similar, although the

  1. Development of an external Faraday cup for beam current measurements

    International Nuclear Information System (INIS)

    Kim, Kye-Ryung; Jung, Myung-Hwan; Ra, Se-Jin; Lee, Seok-Ki

    2010-01-01

    In general, beam current measurements are very important for many kinds of experiments using highly energetic particle beams at accelerators, such as cyclotrons, linacs, etc. The Faraday cup is known to be one of the most popular beam current measurement tools. We developed an external Faraday cup to measure the beam current at a dedicated beam line for low-flux experiments installed at the MC-50 cyclotron of Korea Institute of Radiological and Medical Sciences (KIRAMS). It was designed for external beam current measurements and is composed of a vacuum chamber, an entrance window, a collimator, a electrostatic suppressor ring, and a cup. The window is made of 75-um-thick Kapton film, and the diameter of the collimator is 10 mm or 20 mm. The ring and the cup has 5-cm inner diameters, and the thickness of the bottom of the cup is 2 cm, which is enough to absorb the total proton energy up to 45 MeV. Using this external Faraday cup, we measured the beam current from the cyclotron, and we compared measured flux to the results from film dosimetry using GAF films.

  2. Resistance and sheet resistance measurements using electron beam induced current

    International Nuclear Information System (INIS)

    Czerwinski, A.; Pluska, M.; Ratajczak, J.; Szerling, A.; KaPtcki, J.

    2006-01-01

    A method for measurement of spatially uniform or nonuniform resistance in layers and strips, based on electron beam induced current (EBIC) technique, is described. High electron beam currents are used so that the overall resistance of the measurement circuit affects the EBIC signal. During the evaluation, the electron beam is scanned along the measured object, whose load resistance varies with the distance. The variation is compensated by an adjustable resistance within an external circuit. The method has been experimentally deployed for sheet resistance determination of buried regions of lateral confinements in semiconductor laser heterostructures manufactured by molecular beam epitaxy

  3. Radio-frequency-assisted current startup in the Fusion Engineering Device

    International Nuclear Information System (INIS)

    Borowski, S.K.; Kammash, T.; Martin Peng, Y.K.

    1984-01-01

    Auxiliary radio-frequency (RF) heating of electrons before and during the current rise phase of a large tokamak, such as the Fusion Engineering Device (FED) (R 0 = 4.8 m, a = 1.3 m, sigma = 1.6, B(R 0 ) = 3.62 T), is examined as a means of reducing both the initiation loop voltage and resistive flux expenditure during startup. Prior to current initiation, 1 to 2 MW of electron cyclotron resonance heating power at about90 GHz is used to create a small volume of high conductivity plasma (T /sub e/ approx. = 100 eV, n /sub e/ approx. = 10 19 m -3 ) near the upper hybrid resonance (UHR) region. This plasma conditioning, referred to as preheating, permits a small radius (a 0 approx. = 0.2 to 0.4 m) current channel to be established with a relatively low initial loop voltage (less than or equal to 25 V as opposed to about 100 V without rf assist). During the subsequent plasma expansion and current rise phase, a combination of rf heating (up to 5 MW) and linear current ramping leads to a substantial savings in voltseconds by (a) minimizing the resistive flux consumption and (b) producing broad current density profiles. (With such broad profiles, the internal flux requirements are maintained at or near the flat profile limit.)

  4. The dependence of potential well formation on the magnetic field strength and electron injection current in a polywell device

    International Nuclear Information System (INIS)

    Cornish, S.; Gummersall, D.; Carr, M.; Khachan, J.

    2014-01-01

    A capacitive probe has been used to measure the plasma potential in a polywell device in order to observe the dependence of potential well formation on magnetic field strength, electron injection current, and polywell voltage bias. The effectiveness of the capacitive probe in a high energy electron plasma was determined by measuring the plasma potential of a planar diode with an axial magnetic field. The capacitive probe was translated along the axis of one of the field coils of the polywell, and the spatial profile of the potential well was measured. The confinement time of electrons in the polywell was estimated with a simple analytical model which used the experimentally observed potential well depths, as well as a simulation of the electron trajectories using particle orbit theory

  5. Measurement in Marketing: Current Scenario, Recommendations and Challenges

    Directory of Open Access Journals (Sweden)

    Felipe Zambaldi

    2014-05-01

    Full Text Available The purpose of this article is to discuss about construct measurement in Marketing by summarizing the main considerations about the subject. First, it discusses the origins of the debates about the theme since the 1970s and describes its main consolidated models (the classical Churchill’s model, the COARSE model and the formative measurement model. Then it presents current concerns about the classical approach with relevant recommendations (particularly regarding multi-item measurement, single-item measurement, rating scales and cross-cultural aspects. At the end, it presents considerations about measurement trends in Marketing with emphasis on the Item Response Theory (IRT, Bayesian estimators and Partial Least Squares (PLS. The article updates the debate on the theme and contributes to Marketing experts and researchers who demand a current view about measurement and recommendations for research development. 

  6. Migrational polarization in high-current density molten salt electrochemical devices

    Energy Technology Data Exchange (ETDEWEB)

    Braunstein, J.; Vallet, C.E.

    1977-01-01

    Electrochemical flux equations based on the thermodynamics of irreversible processes have been derived in terms of experimental transport coefficients for binary molten salt mixtures analogous to those proposed for high temperature batteries and fuel cells. The equations and some numerical solutions indicate steady state composition gradients of significant magnitude. The effects of migrational separation must be considered along with other melt properties in the characterization of electrode behavior, melt composition, operating temperatures and differences of phase stability, wettability and other physicochemical properties at positive and negative electrodes of high current density devices with mixed electrolytes.

  7. Analysis of minor disruptions during current flat phase in the HL-1 device

    International Nuclear Information System (INIS)

    Yan Longwen; Shi Bingren; Zheng Yongzhen; Peng Liling; Huang Keqiang

    1991-01-01

    The phenomena of minor disruptions during current flat phase have been observed in the HL-1 device for five years. When these phenomena appear the safety factors qa are between 2.5∼4.5, and densities are between 1 x 10 13 cm -3 and 3 x 10 13 cm -3 or greater than 4 x 10 13 cm -3 . Periodic relaxation ocsillations of voltage and soft-X-rays are observed during minor disruptions. Their character and development conditions are carefully analylsed

  8. Understanding quantum mechanics by measuring the properties of mesoscopic devices

    International Nuclear Information System (INIS)

    Webb, R.

    1993-01-01

    Measurements of the electrical transport and magnetic properties of micron-size scale insulators, metals, semi-metals, and semiconductors at low temperatures have uncovered a wealth of unexpected phenomena. The only way to understand these new properties is by invoking many of the postulates of quantum mechanics. The author has confirmed that the electron acts as a long-range phase-coherent wave and conventional classical forces are not as important as scalar and vector potentials in determining the response of the electron as it moves through its environment. This talk will focus on the measurement of the Aharonov-Bohm self-interference effects, nonlocal transport phenomena, and persistent currents in normal metal ring structures that have been observed in these nanostructures

  9. Analyses on the measurement of leakage currents in CdZnTe radiation detectors

    International Nuclear Information System (INIS)

    Mescher, M.J.; Hoburg, J.F.; Schlesinger, T.E.; James, R.B.

    1999-01-01

    Models that place design constraints on devices which are used to measure the leakage currents in high-resistivity semiconductor materials are presented. If these design constraints are met, these models can then be used to quantitatively predict the surface sheet resistance of devices which are dominated by surface leakage currents. As a result, a means is provided to directly compare passivation techniques which are developed to decrease surface leakage currents. Furthermore, these models illustrate the necessity for inclusion of relevant geometrical data on sample size and shape and electrode configuration when reporting results of surface passivation techniques. These models specifically examine the case where a dc potential is applied across two electrodes on the surface of a semiconductor substrate which has a surface layer with lower resistivity than the bulk material. The authors describe several of the more common configurations used in analyzing passivation techniques for compounds of Cd 1-x Zn x Te (CZT) used for room-temperature radiation detection

  10. Current Measurements and Overwash Monitoring Using Tilt Current Meters in Three Coastal Environments

    Science.gov (United States)

    Lowell, N. S.; Sherwood, C. R.; Decarlo, T. M.; Grant, J. R.

    2014-12-01

    Tilt Current Meters (TCMs) provide accurate, cost effective measurements of near-bottom current velocities. Many studies in coastal environments require current measurements, which are frequently made with Acoustic Doppler Profilers (ADPs). ADPs are expensive, however, and may not be suitable for locations where there is significant risk of damage, loss, or theft or where a large spatial array of measurements is required. TCMs, by contrast, are smaller, less expensive, and easier to deploy. This study tested TCMs in three sites to determine their suitability for use in research applications. TCMs are based on the drag-tilt principle, where the instrument tilts in response to current. The meter consists of a buoyant float with an onboard accelerometer, three-axis tilt sensor, three-axis magnetometer (compass), and a data logger. Current measurements are derived by post processing the tilt and compass values and converting them to velocity using empirical calibration data. Large data-storage capacity (4 GB) and low power requirements allow long deployments (many months) at high sample rates (16 Hz). We demonstrate the utility of TCM current measurements on a reef at Dongsha Atoll in the South China Sea, and in Vineyard Sound off Cape Cod, where the TCM performance was evaluated against ADP measurements. We have also used the TCM to record waves during an overwash event on a Cape Cod barrier beach during a winter storm. The TCM recorded waves as they came through the overwash channel, and the data were in agreement with the water-level record used as a reference. These tests demonstrate that TCMs may be used in a variety of near shore environments and have the potential to significantly increase the density of meters in future studies were current measurements are required.

  11. 77 FR 35745 - Highway Safety Programs; Conforming Products List of Screening Devices To Measure Alcohol in...

    Science.gov (United States)

    2012-06-14

    ..., battery powered device with a semiconductor sensor. (2) Alcohol Countermeasure Systems Corp., submitted...-0062] Highway Safety Programs; Conforming Products List of Screening Devices To Measure Alcohol in... Screening Devices to Measure Alcohol in Bodily Fluids dated, March 31, 2008 (73 FR 16956). DATES: Effective...

  12. Measuring device for the temperature coefficient of reactor moderators

    International Nuclear Information System (INIS)

    Nakano, Yuzo.

    1987-01-01

    Purpose: To rapidly determine by automatic calculation the temperature coefficient for moderators which has been determined so far by a log of manual processings. Constitution: Each of signals from a control rod position indicator, a reactor reactivity, instrument and moderator temperature meter are inputted, and each of the signals and designed valued for the doppler temperature coefficients are stored. Recurling calculation is conducted based on the reactivity and the moderator temperature at an interval where the temperature changes of the moderators are equalized at an identical control rod position, to determine isothermic coefficient. Then, the temperature coefficient for moderator are calculated from the isothermic coefficient and the doppler temperature coefficient. The relationship between the reactivity and the moderator temperature is plotted on a X-Y recorder. The stored signals and the calculated temperature coefficient for moderators are sequentially displayed and the results are printed out when the measurement is completed. According to the present device, since the real time processing is conducted, the processing time can be shortened remarkably. Accordingly, it is possible to save the man power for the test of the nuclear reactor and improve the reactor operation performance. (Kamimura, M.)

  13. The Development of 1Balance: A Connected Medical Device for Measuring Human Balance

    Directory of Open Access Journals (Sweden)

    Heikki Sjöman

    2018-05-01

    Full Text Available Prototyping (iterative loops of design–build–test is a proven method of efficiently developing new products. Developing products not only quickly, but that are also fit for purpose, implies engaging the end users and iterating the technology at hand. However, there is currently little research on how engineering design can approach developing connected devices. The purpose of this paper is to distinguish and discuss design approaches that are suitable for connected devices. Internet of Things devices consist of both the physical products themselves and the data that is coming out of the products, which we define as the external and internal data, respectively. They both can be prototyped separately, but since the data acquired can influence the design of the device and vice versa, we propose to link these two together in the product development process. This issue becomes more apparent when designing networks of sensors, e.g., for complex artificial intelligence (AI databases. We explain the principle by describing the development of 1Balance through six different prototypes for human balance measurement. Technologically quantifying balance is an underused approach for objectively evaluating the state of a human’s performance. The authors have developed a mobile application for monitoring balance as a physiological signal (amount of sway via a compact wireless inertial measurement unit (IMU sensor strapped to the body of the subject for the duration of the measurement. We describe the design process for developing this connected medical device, as well as how the acquired data was used to improve the design of the product. In conclusion, we propose conceptually connecting the external and internal data prototyping loops.

  14. A measurement of perpendicular current density in an aurora

    International Nuclear Information System (INIS)

    Bering, E.A.; Mozer, F.S.

    1975-01-01

    A Nike Tomahawk sounding rocket was launched into a 400-γ auroral substorm on February 7, 1972, from Esrange, Kiruna, Sweden. The rocket instrumentation included a split Langmuir probe plasma velocity detector and a double-probe electric field detector. Above 140-km altitude the electric field deduced from the ion flow velocity measurement and the electric field measured by the double probe agree to an accuracy within the uncertainties of the two measurements. The difference between the two measurements at altitudes below 140 km provides an in situ measurement of current density and conductivity. Alternatively, if values for the conductivity are assumed, the neutral wind velocity can be deduced. The height-integrated current was 0.11 A/m flowing at an azimuth of 276degree. The neutral winds were strong, exhibited substantial altitude variation in the east-west component, and were predominantly southward

  15. Current measurements by Faraday rotation in single mode optical fibers

    International Nuclear Information System (INIS)

    Chandler, G.I.; Jahoda, F.C.

    1984-01-01

    Development of techniques for measuring magnetic fields and currents by Faraday rotation in single-mode optical fibers has continued. We summarize the results of attempts to measure the toroidal plasma current in the ZT-40 Reversed-Field-Pinch using multi-turn fiber coils. The fiber response is reproducible and in accord with theory, but the amount and distribution of the stress-induced birefringence in this case are such that prediction of the sensor response at low currents is difficult if not impossible. The low-current difficulty can be overcome by twisting the fiber to induce a circular birefringence bias. We report the results of auxiliary experiments with a fiber that has been twisted with 15 turns per meter and then re-coated to lock the twist in place

  16. Automatic control system for measuring currents produced by ionization chambers

    International Nuclear Information System (INIS)

    Brancaccio, Franco

    2002-01-01

    Ionization Chambers in current mode operation are usually used in Nuclear Metrology. Activity measurements are quickly performed by Ionization Chambers, with very good precision. For this purpose measurements of very low ionization currents, carried out by high quality instrumentation, are required. Usually, electrometers perform the current integration method under command of signals from an automation system, in order to reduce the measurement uncertainties. Among the measurement systems at the Laboratorio de Metrologia Nuclear (LMN) of IPEN, there are two ionization chamber systems. In the present work, an automation system developed for current integration measurements is described. This automation system is composed by software (graphic interface and control) and an electronic module connected to a microcomputer, by means of a commercial data acquisition card. Several test measurements were performed in order to determine the intrinsic uncertainty, linearity and stability of the system. Using calibrated radioactive solutions, the IG12/A20 chamber calibration factors for 18 F and 153 Sm were obtained, making possible to determine activities of these radionuclides. (author)

  17. Lightning Current Measurement with Fiber-Optic Sensor

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George N.; Mata, Carlos T.; Mata, Angel G.; Snyder, Gary P.

    2014-01-01

    A fiber-optic current sensor is successfully developed with many potential applications for electric current measurement. Originally developed for in-flight lightning measurement, the sensor utilizes Faraday Effect in an optical fiber. The Faraday Effect causes linear light polarization in a fiber to rotate when the fiber is exposed to a magnetic field. The polarization change is detected using a reflective polarimetric scheme. Forming fiber loops and applying Ampere's law, measuring the total light rotation results in the determination of the total current enclosed. The sensor is conformable to complex structure geometry. It is also non-conductive and immune to electromagnetic interference, saturation or hysteresis. Installation is non-intrusive, and the sensor can be safely routed through flammable areas. Two similar sensor systems are described in this paper. The first system operates at 1310nm laser wavelength and is capable of measuring approximately 300 A - 300 kA, a 60 dB range. Laboratory validation results of aircraft lighting direct and in-direct effect current amplitudes are reported for this sensor. The second system operates at 1550nm wavelength and can measure about 400 A - 400 kA. Triggered-lightning measurement data are presented for this system. Good results are achieved in all cases.

  18. Proton current measurements using the prompt gamma ray diagnostic technique

    International Nuclear Information System (INIS)

    Leeper, R.J.; Burns, E.J.T.; Johnson, D.J.; McMurtry, W.M.

    1981-01-01

    Prompt gamma ray signals from the nuclear reaction 7 Li(p,γ) 8 Be have been used to make time resolved proton current measurements. In these measurements, the proton beam was allowed to strike cylindrical thick lithium metal targets. The time integrated proton current was measured using gamma activation of copper via the reaction 63 Cu(γ,n) 62 Cu(β+). The positron activity of the copper sample was easily measured using coincidence counting techniques. The number of 62 Cu atoms produced per proton incident on a thick Li metal target was determined with separate calibration runs performed on the Sandia 2.5 MeV Van de Graaff accelerator. The time history of the prompt gamma production was measured using six EGG NPM-54 scintillator photomultiplier combinations shielded by 96.5 cm of concrete and 5.1 cm of Pb. The use of six scintillator photomultiplier combinations was necessary to increase the statistical precision of the data. The normalization of the prompt gamma time history data with the total time integrated proton-current measurement yielded the absolute time resolved proton current on target. Data from runs performed on the Sandia Proto I accelerator will be presented

  19. Surface Passivation for Reliable Measurement of Bulk Electronic Properties of Heterojunction Devices.

    Science.gov (United States)

    Bissig, Benjamin; Guerra-Nunez, Carlos; Carron, Romain; Nishiwaki, Shiro; La Mattina, Fabio; Pianezzi, Fabian; Losio, Paolo A; Avancini, Enrico; Reinhard, Patrick; Haass, Stefan G; Lingg, Martina; Feurer, Thomas; Utke, Ivo; Buecheler, Stephan; Tiwari, Ayodhya N

    2016-10-01

    Quantum efficiency measurements of state of the art Cu(In,Ga)Se 2 (CIGS) thin film solar cells reveal current losses in the near infrared spectral region. These losses can be ascribed to inadequate optical absorption or poor collection of photogenerated charge carriers. Insight on the limiting mechanism is crucial for the development of more efficient devices. The electron beam induced current measurement technique applied on device cross-sections promises an experimental access to depth resolved information about the charge carrier collection probability. Here, this technique is used to show that charge carrier collection in CIGS deposited by multistage co-evaporation at low temperature is efficient over the optically active region and collection losses are minor as compared to the optical ones. Implications on the favorable absorber design are discussed. Furthermore, it is observed that the measurement is strongly affected by cross-section surface recombination and an accurate determination of the collection efficiency is not possible. Therefore it is proposed and shown that the use of an Al 2 O 3 layer deposited onto the cleaved cross-section significantly improves the accuracy of the measurement by reducing the surface recombination. A model for the passivation mechanism is presented and the passivation concept is extended to other solar cell technologies such as CdTe and Cu 2 (Zn,Sn)(S,Se) 4 . © 2016 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  20. Physiotherapy and cardiac rhythm devices: a review of the current scope of practice.

    Science.gov (United States)

    Digby, Geneviève C; Daubney, Marguerite E; Baggs, Jim; Campbell, Debra; Simpson, Christopher S; Redfearn, Damian P; Brennan, F James; Abdollah, Hoshiar; Baranchuk, Adrian

    2009-07-01

    Several case reports have demonstrated negative interactions between various physiotherapy modalities and cardiac rhythm devices (CRD). Fear of these potential interactions may lead to suboptimal utilization of physiotherapy treatments in CRD patients. No prior review of available guidelines, or management strategies, on the interaction between physiotherapy modalities and CRD patients has been reported. To review existing guidelines regarding the use of physiotherapy modalities in patients with pacemakers and/or implantable cardioverter-defibrillators (ICDs). To retrospectively analyse CRD patient encounters at a local physiotherapy facility during a period of 2 years. A review of the literature regarding the potential interactions between physiotherapy modalities and CRDs was performed. Next, a 2 year retrospective analysis of patient encounters at a physiotherapy facility was conducted. In addition, seven international physiotherapy societies and four CRD manufacturers were surveyed with respect to recommendations regarding physiotherapy treatments in device patients. The local physiotherapy facility treated 25 patients with CRD (22 pacemaker and 3 ICD patients) for a total of 230 visits (9.2 visits/patient). Five patients received transcutaneous electrical nerve stimulation (TENS) and all 25 were administered additional treatment in the form of ultrasound (15), acupuncture (19), Laser (7), traction/manual therapy (12), exercise (8), education (18), taping (5), and/or moist heat (5). No complications occurred. Meanwhile, international societies and device manufacturers offered few specific or consistent recommendations. There are no specific international policies regarding the administration of physiotherapy modalities in CRD patients and, thus, there are no specific guidelines to be implemented at the local level. Review of the literature and of recommendations from CRD manufacturers suggests that TENS, Diathermy, and Interferential Electrical Current Therapy

  1. Determination of current and rotational transform profiles in a current-carrying stellarator using soft x-ray emissivity measurements

    Science.gov (United States)

    Ma, X.; Cianciosa, M. R.; Ennis, D. A.; Hanson, J. D.; Hartwell, G. J.; Herfindal, J. L.; Howell, E. C.; Knowlton, S. F.; Maurer, D. A.; Traverso, P. J.

    2018-01-01

    Collimated soft X-ray (SXR) emissivity measurements from multi-channel cameras on the Compact Toroidal Hybrid (CTH) tokamak/torsatron device are incorporated in the 3D equilibrium reconstruction code V3FIT to reconstruct the shape of flux surfaces and infer the current distribution within the plasma. Equilibrium reconstructions of sawtoothing plasmas that use data from both SXR and external magnetic diagnostics show the central safety factor to be near unity under the assumption that SXR iso-emissivity contours lie on magnetic flux surfaces. The reconstruction results are consistent with those using the external magnetic data and a constraint on the location of q = 1 surfaces determined from the sawtooth inversion surface extracted from SXR brightness profiles. The agreement justifies the use of approximating SXR emission as a flux function in CTH, at least within the core of the plasma, subject to the spatial resolution of the SXR diagnostics. This improved reconstruction of the central current density indicates that the current profile peakedness decreases with increasing external transform and that the internal inductance is not a relevant measure of how peaked the current profile is in hybrid discharges.

  2. Measurement of fast risetime megampere currents by quartz gauge

    International Nuclear Information System (INIS)

    Williams, R.R.; McDaniel, D.H.; Stinnett, R.W.

    1980-01-01

    Quartz gauges have been used on the Sandia National Laboratories Proto II accelerator to measure current in the magnetically insulated transmission line at the 11 TW power level. The accelerator delivers 3.5 MA at 2 x 10 14 A/s in a 40 ns pulse to a 0.0127 m diameter aluminum liner to produce a high density plasma. At this radius and dI/dt levels, the B-dot monitors no longer function for the measurement of load current because the monitor suffers electrical breakdown. Quartz pressure gauges mounted at a radius of 0.0086 m have successfully measured the magnetic pressure due to the load current with nanosecond temporal resolution

  3. Simulation of vibration-induced effect on plasma current measurement using a fiber optic current sensor.

    Science.gov (United States)

    Descamps, Frédéric; Aerssens, Matthieu; Gusarov, Andrei; Mégret, Patrice; Massaut, Vincent; Wuilpart, Marc

    2014-06-16

    An accurate measurement of the plasma current is of paramount importance for controlling the plasma magnetic equilibrium in tokamaks. Fiber optic current sensor (FOCS) technology is expected to be implemented to perform this task in ITER. However, during ITER operation, the vessel and the sensing fiber will be subject to vibrations and thus to time-dependent parasitic birefringence, which may significantly compromise the FOCS performance. In this paper we investigate the effects of vibrations on the plasma current measurement accuracy under ITER-relevant conditions. The simulation results show that in the case of a FOCS reflection scheme including a spun fiber and a Faraday mirror, the error induced by the vibrations is acceptable regarding the ITER current diagnostics requirements.

  4. Current Consumption Measurements with a Carrier Aggregation Smartphone

    DEFF Research Database (Denmark)

    Sanchez-Mejias, Rafael; Guo, Yu; Lauridsen, Mads

    2014-01-01

    on this topic. Currently there are only theoretical expansions of LTE release 8 power models available, but this article presents the first publicly available current consumption measurements on a commercial CA-capable UE. In this work it is examined how the activation and use of CA (10 + 10 MHz) affects the UE......Carrier Aggregation (CA) is introduced in LTE release 10 to improve data rates by allowing the User Equipment (UE) to receive data on more than one LTE carrier. The related increased complexity is expected to affect the UE current consumption, but yet no empirical evaluation has been published...

  5. A device for fresh water sampling before radioactive measurements

    International Nuclear Information System (INIS)

    Maubert, Henri; Picat, Philippe.

    1982-06-01

    On account of the many field operations carried out by the laboratory, a water sampling device has been developed. This portable autonomous device performs in situ water filtration and concentration on ion exchange resins and activated carbon columns. The device is described and the trapping performance for 8 radionuclides is given. A comparison is made with the so-called evaporation method. The effects of the treatment of the filtrating elements on the radioactive results are studied. This sampling method is very sensitive [fr

  6. Measuring electric conductivity in liquid metals by eddy current method

    International Nuclear Information System (INIS)

    Zhuravlev, S.P.; Ostrovskij, O.I.; Grigoryan, V.A.

    1982-01-01

    Technique permitting to apply the method of vertiginous currents for investigation of electric conductivity of metal melts in the high temperature range is presented. Interferences affecting accuracy of measurements are specified and ways of their removing are pointed out. Scheme of measuring and design of the facility are described. Results of measuring electric resistance of liquid Fe, Co, Ni obtained for the first time by this method are presented. The data obtained agree with the results of measurements conducted by the method of the rotating magnetic field. Difference in absolute values of electric resistance in parallel experiments for each metal does not exceed 4%

  7. Determining Confounding Sensitivities In Eddy Current Thin Film Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Gros, Ethan; Udpa, Lalita; Smith, James A.; Wachs, Katelyn

    2016-07-01

    Determining Confounding Sensitivities In Eddy Current Thin Film Measurements Ethan Gros, Lalita Udpa, Electrical Engineering, Michigan State University, East Lansing MI 48824 James A. Smith, Experiment Analysis, Idaho National Laboratory, Idaho Falls ID 83415 Eddy current (EC) techniques are widely used in industry to measure the thickness of non-conductive films on a metal substrate. This is done using a system whereby a coil carrying a high-frequency alternating current is used to create an alternating magnetic field at the surface of the instrument's probe. When the probe is brought near a conductive surface, the alternating magnetic field will induce ECs in the conductor. The substrate characteristics and the distance of the probe from the substrate (the coating thickness) affect the magnitude of the ECs. The induced currents load the probe coil affecting the terminal impedance of the coil. The measured probe impedance is related to the lift off between coil and conductor as well as conductivity of the test sample. For a known conductivity sample, the probe impedance can be converted into an equivalent film thickness value. The EC measurement can be confounded by a number of measurement parameters. It is the goal of this research to determine which physical properties of the measurement set-up and sample can adversely affect the thickness measurement. The eddy current testing is performed using a commercially available, hand held eddy current probe (ETA3.3H spring loaded eddy probe running at 8 MHz) that comes with a stand to hold the probe. The stand holds the probe and adjusts the probe on the z-axis to help position the probe in the correct area as well as make precise measurements. The signal from the probe is sent to a hand held readout, where the results are recorded directly in terms of liftoff or film thickness. Understanding the effect of certain factors on the measurements of film thickness, will help to evaluate how accurate the ETA3.3H spring

  8. Design of a device for simultaneous particle size and electrostatic charge measurement of inhalation drugs.

    Science.gov (United States)

    Zhu, Kewu; Ng, Wai Kiong; Shen, Shoucang; Tan, Reginald B H; Heng, Paul W S

    2008-11-01

    To develop a device for simultaneous measurement of particle aerodynamic diameter and electrostatic charge of inhalation aerosols. An integrated system consisting of an add-on charge measurement device and a liquid impinger was developed to simultaneously determine particle aerodynamic diameter and electrostatic charge. The accuracy in charge measurement and fine particle fraction characterization of the new system was evaluated. The integrated system was then applied to analyze the electrostatic charges of a DPI formulation composed of salbutamol sulphate-Inhalac 230 dispersed using a Rotahaler. The charge measurement accuracy was comparable with the Faraday cage method, and incorporation of the charge measurement module had no effect on the performance of the liquid impinger. Salbutamol sulphate carried negative charges while the net charge of Inhalac 230 and un-dispersed salbutamol sulphate was found to be positive after being aerosolized from the inhaler. The instantaneous current signal was strong with small noise to signal ratio, and good reproducibility of charge to mass ratio was obtained for the DPI system investigated. A system for simultaneously measuring particle aerodynamic diameter and aerosol electrostatic charges has been developed, and the system provides a non-intrusive and reliable electrostatic charge characterization method for inhalation dosage forms.

  9. Cryogenic Current Comparator for Absolute Measurement of the Dark Current of the Superconducting Cavities for Tesla

    CERN Document Server

    Knaack, K; Wittenburg, K

    2003-01-01

    A newly high performance SQUID based measurement system for detecting dark currents, generated by superconducting cavities for TESLA is proposed. It makes use of the Cryogenic Current Comparator principle and senses dark currents in the nA range with a small signal bandwidth of 70 kHz. To reach the maximum possible energy in the TESLA project is a strong motivation to push the gradients of the superconducting cavities closer to the physical limit of 50 MV/m. The field emission of electrons (the so called dark current) of the superconducting cavities at strong fields may limit the maximum gradient. The absolute measurement of the dark current in correlation with the gradient will give a proper value to compare and classify the cavities. This contribution describes a Cryogenic Current Comparator (CCC) as an excellent and useful tool for this purpose. The most important component of the CCC is a high performance DC SQUID system which is able to measure extremely low magnetic fields, e.g. caused by the extracted ...

  10. Application possibility of the direct current conduction method for nondestructive crack measurement

    International Nuclear Information System (INIS)

    Riedl, R.

    1982-01-01

    An important value to determine the danger of cracks is the determination of crack depths. The crack depth can be determined quite accurate by means of the direct current conduction method, if one holds onto certain rules. Often complicated experimental set-ups are applied. However, portable commercial devices can be obtained that can be used for partial fluxation, that yield good results. By means of two examples: crack conduction samples in which the built-up of a constant-cracking is persued up to a certain depth, as well as the persuasion of an continuing crack in a bearing cylinder, shall be demonstrated that is very well possible to record accurate profiles with commercial devices and to avoid expensive measurement devices. (orig.) [de

  11. Investigation of disorder and its effect on electrical transport in electrochemically doped polymer devices by current-voltage and impedance spectroscopy

    Science.gov (United States)

    Rahman Khan, Motiur; Anjaneyulu, P.; Koteswara Rao, K. S. R.; Menon, R.

    2017-03-01

    We report on the analysis of temperature-dependent current-voltage characteristics and impedance measurements of electrochemically doped poly(3-methylthiophene) devices at different doping levels. The extent of doping is carefully tailored such that only the bulk-limited transport mechanism prevails. A transition from exponentially distributed trap-limited transport to trap-free space-charge-limited current is observed in current-voltage conduction upon increasing the doping. The obtained trap densities (3.2  ×  1016 cm-3 and 8.6  ×  1015 cm-3) and trap energies (31.7 meV and 16.6 meV) for different devices signify the variation in disorder with doping, which is later supported by impedance measurements. Impedance-frequency data for various devices can not be explained using the parallel resistance-capacitance (RC) model in the equivalent circuit. However, this was established by incorporating a constant phase element Q (CPE) instead of the capacitance parameter. It should be emphasized that low doping devices in particular are best simulated with two CPE elements, while the data related to other devices are fitted well with a single CPE element. It is also observed from evaluated circuit parameters that the spatial inhomogeneity and disorder are the cause of variability in different samples, which has an excellent correlation with the temperature-dependent current-voltage characteristics.

  12. Comparison of current tonometry techniques in measurement of intraocular pressure.

    Science.gov (United States)

    Kouchaki, Behrooz; Hashemi, Hassan; Yekta, Abbasali; Khabazkhoob, Mehdi

    2017-06-01

    To compare four tonometry techniques: Goldmann applanation tonometer (GAT), Dynamic contour tonometer (DCT), Non-contact tonometer (NCT), and Ocular Response Analyzer (ORA) in the measurement of intraocular pressure (IOP) and the impact of some corneal biomechanical factors on their performance. In this cross-sectional study, volunteers with normal ophthalmic examination and no history of eye surgery (except for uncomplicated cataract surgery) or trauma were selected. Twenty-five subjects were male, and 21 were female. The mean age was 48 ± 19.2 years. Anterior segment parameters were measured with Scheimpflug imaging. IOP was measured with GAT, DCT, NCT, and ORA in random order. A 95% limit of agreement of IOPs was analyzed. The impact of different parameters on the measured IOP with each device was evaluated by regression analysis. The average IOP measured with GAT, DCT, NCT, and ORA was 16.4 ± 3.5, 18.1 ± 3.4, 16.2 ± 3.9, and 17.3 ± 3.4 mmHg, respectively. The difference of IOP measured with NCT and GAT was not significant ( P  = 0.382). Intraocular pressure was significantly different between GAT with DCT and IOP CC ( P  tonometers.

  13. A fiber-optic current sensor for lightning measurement applications

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George N.

    2015-05-01

    An optical-fiber sensor based on Faraday Effect is developed for measuring total lightning electric current. It has many unique capabilities not possible with traditional current sensors. Designed for aircraft installation, the sensor is lightweight, non-conducting, structure-conforming, and is immune to electromagnetic interference, hysteresis and saturation. It can also be used on windmills, lightning towers, and can help validate lightning detection network measurements. Faraday Effect causes light polarization to rotate when the fiber is exposed to a magnetic field in the direction of light propagation. Thus, the magnetic field strength can be determined from the light polarization change. By forming closed fiber loops and applying Ampere's law, measuring the total light rotation yields the total current enclosed. The broadband, dual-detector, reflective polarimetric scheme allows measurement of both DC component and AC waveforms with about 60 dB dynamic range. Three sensor systems were built with different sensitivities from different laser wavelengths. Operating at 850nm, the first system uses twisted single-mode fiber and has a 150 A - 150 KA range. The second system operates at 1550nm, uses spun polarization maintaining fiber, and can measure 400 A - 400 KA. Both systems were validated with rocket-triggered lightning measurements and achieved excellent results when compared to a resistive shunt. The third system operates at 1310nm, uses spun polarization maintaining fiber, and can measure approximately 300 A - 300 KA. High current measurements up to 200 KA were demonstrated at a commercial lightning test facility. The system was recently installed on an aircraft and flown near icing weather conditions.

  14. A Fiber-Optic Current Sensor for Lightning Measurement Applications

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George N.

    2015-01-01

    An optical-fiber sensor based on Faraday Effect is developed for measuring total lightning electric current. It has many unique capabilities not possible with traditional current sensors. Designed for aircraft installation, the sensor is lightweight, non-conducting, structure-conforming, and is immune to electromagnetic interference, hysteresis and saturation. It can also be used on windmills, lightning towers, and can help validate lightning detection network measurements. Faraday Effect causes light polarization to rotate when the fiber is exposed to a magnetic field in the direction of light propagation. Thus, the magnetic field strength can be determined from the light polarization change. By forming closed fiber loops and applying Ampere's law, measuring the total light rotation yields the total current enclosed. The broadband, dual-detector, reflective polarimetric scheme allows measurement of both DC component and AC waveforms with about 60 dB dynamic range. Three sensor systems were built with different sensitivities from different laser wavelengths. Operating at 850nm, the first system uses twisted single-mode fiber and has a 150 A - 150 KA range. The second system operates at 1550nm, uses spun polarization maintaining fiber, and can measure 400 A - 400 KA. Both systems were validated with rocket-triggered lightning measurements and achieved excellent results when compared to a resistive shunt. The third system operates at 1310nm, uses spun polarization maintaining fiber, and can measure approximately 300 A - 300 KA. High current measurements up to 200 KA were demonstrated at a commercial lightning test facility. The system was recently installed on an aircraft and flown near icing weather conditions.

  15. Model for an irreversible bias current in the superconducting qubit measurement process

    International Nuclear Information System (INIS)

    Hutchinson, G. D.; Williams, D. A.; Holmes, C. A.; Stace, T. M.; Spiller, T. P.; Barrett, S. D.; Milburn, G. J.; Hasko, D. G.

    2006-01-01

    The superconducting charge-phase ''quantronium'' qubit is considered in order to develop a model for the measurement process used in the experiment of Vion et al. [Science 296, 886 (2002)]. For this model we propose a method for including the bias current in the readout process in a fundamentally irreversible way, which to first order is approximated by the Josephson junction tilted-washboard potential phenomenology. The decohering bias current is introduced in the form of a Lindblad operator and the Wigner function for the current-biased readout Josephson junction is derived and analyzed. During the readout current pulse used in the quantronium experiment we find that the coherence of the qubit initially prepared in a symmetric superposition state is lost at a time of 0.2 ns after the bias current pulse has been applied, a time scale that is much shorter than the experimental readout time. Additionally we look at the effect of Johnson-Nyquist noise with zero mean from the current source during the qubit manipulation and show that the decoherence due to the irreversible bias current description is an order of magnitude smaller than that found through adding noise to the reversible tilted-washboard potential model. Our irreversible bias current model is also applicable to persistent-current-based qubits where the state is measured according to its flux via a small-inductance direct-current superconducting quantum interference device

  16. Surface Current Measurements In Terra Nova Bay By Hf Radar

    Science.gov (United States)

    Flocco, D.; Falco, P.; Wadhams, P.; Spezie, G.

    We present the preliminary results of a field experiment carried out within frame- work of the CLIMA project of the Italian National Programme for Antarctic Research (PNRA) and in cooperation with the Scott Polar Research Institute of Cambridge. Dur- ing the second period (02/12/1999-23/01/2000) of the XV Italian expedition a coastal radar was used to characterize the current field in the area of Terra Nova Bay (TNB). One of the aims of the CLIMA (Climatic Long-term Interactions for the Mass balance in Antarctica) project is to determine the role of the polynya in the sea ice mass bal- ance, water structure and local climate. The OSCR-II experiment was planned in order to provide surface current measurements in the area of TNB polynya, one of the most important coastal polynya of the Ross Sea. OSCR (Ocean Surface Current Radar) is a shore based, remote sensing system designed to measure sea surface currents in coastal waters. Two radar sites (a master and a slave) provide with radial current mea- surements; data combined from both sites yield the total current vector. Unfortunately the master and slave stations did not work together throughout the whole period of the experiment. A description of the experiment and a discussion of the results, will be proposed.

  17. The development of an enhanced strain measurement device to support testing of radioactive material packages

    International Nuclear Information System (INIS)

    Uncapkher, W.L.; Arviso, M.

    1995-01-01

    Radioactive material package designers use structural testing to verify and demonstrate package performance. A major part of evaluating structural response is the collection of reliable instrumentation measurement data. Over the last four decades, Sandia National Laboratories (SNL) has been actively involved in the development, testing, and evaluation of measurement devices for a broad range of applications, resulting in the commercialization of several measurement devices commonly used today. SNL maintains an ongoing program sponsored by the US Department of Energy (DOE) to develop and evaluate measurement devices to support testing of packages used to transport radioactive or hazardous materials. The development of the enhanced strain measurement device is part of this program

  18. Direct observation of the leakage current in epitaxial diamond Schottky barrier devices by conductive-probe atomic force microscopy and Raman imaging

    Science.gov (United States)

    Alvarez, J.; Boutchich, M.; Kleider, J. P.; Teraji, T.; Koide, Y.

    2014-09-01

    The origin of the high leakage current measured in several vertical-type diamond Schottky devices is conjointly investigated by conducting probe atomic force microscopy and confocal micro-Raman/photoluminescence imaging analysis. Local areas characterized by a strong decrease of the local resistance (5-6 orders of magnitude drop) with respect to their close surrounding have been identified in several different regions of the sample surface. The same local areas, also referenced as electrical hot-spots, reveal a slightly constrained diamond lattice and three dominant Raman bands in the low-wavenumber region (590, 914 and 1040 cm-1). These latter bands are usually assigned to the vibrational modes involving boron impurities and its possible complexes that can electrically act as traps for charge carriers. Local current-voltage measurements performed at the hot-spots point out a trap-filled-limited current as the main conduction mechanism favouring the leakage current in the Schottky devices.

  19. Fiber-Optic Current Sensor Validation with Triggered Lightning Measurements

    Science.gov (United States)

    Nguyen, Truong X.; Ely, Jay J.; Szatkowski, George N.; Mata, Carlos T.; Mata, Angel G.; Snyder, Gary P.

    2013-01-01

    A fiber optic current sensor based on the Faraday Effect is developed that is highly suitable for aircraft installation and can measure total current enclosed in a fiber loop down to DC. Other attributes include being small, light-weight, non-conducting, safe from electromagnetic interference, and free of hysteresis and saturation. The Faraday Effect causes light polarization to rotate when exposed to a magnetic field in the direction of light propagation. Measuring the induced light polarization rotation in fiber loops yields the total current enclosed. Two sensor systems were constructed and installed at Camp Blanding, Florida, measuring rocket-triggered lightning. The systems were similar in design but with different laser wavelengths, sensitivities and ranges. Results are compared to a shunt resistor as reference. The 850nm wavelength system tested in summer 2011 showed good result comparison early. However, later results showed gradual amplitude increase with time, attributed to corroded connections affecting the 50-ohm output termination. The 1550nm system also yielded good results in the summer 2012. The successful measurements demonstrate the fiber optic sensor's accuracies in capturing real lightning currents, and represent an important step toward future aircraft installation.

  20. SQUID Based Cryogenic Current Comparator for Measurements of the Dark Current of Superconducting Cavities

    CERN Document Server

    Vodel, W; Neubert, R; Nietzsche, S

    2005-01-01

    This contribution presents a LTS-SQUID based Cryogenic Current Comparator (CCC) for detecting dark currents, generated e.g. by superconducting cavities for the upcoming X-FEL project at DESY. To achieve the maximum possible energy the gradients of the superconducting RF cavities should be pushed close to the physical limit of 50 MV/m. The measurement of the undesired field emission of electrons (the so-called dark current) in correlation with the gradient will give a proper value to compare and classify the cavities. The main component of the CCC is a high performance LTS-DC SQUID system which is able to measure extremely low magnetic fields, e.g. caused by the extracted dark current. For this reason the input coil of the SQUID is connected across a special designed toroidal niobium pick-up coil (inner diameter: about 100 mm) for the passing electron beam. A noise limited current resolution of nearly 2 pA/√(Hz) with a measurement bandwidth of up to 70 kHz was achieved without the pick-up coil. Now, ...

  1. An accurate low current measurement circuit for heavy iron beam current monitor

    International Nuclear Information System (INIS)

    Zhou Chaoyang; Su Hong; Mao Ruishi; Dong Chengfu; Qian Yi; Kong Jie

    2012-01-01

    Heavy-ion beams at 10 6 particles per second have been applied to the treatment of deep-seated inoperable tumors in the therapy terminal of the Heavy Ion Research Facility in Lanzhou (HIRFL) which is located at the Institute of Modern Physics, Chinese Academy of Sciences (IMP, CAS). An accurate low current measurement circuit following a Faraday cup was developed to monitor the beam current at pA range. The circuit consisted of a picoammeter with a bandwidth of 1 kHz and a gated integrator (GI). A low input bias current precision amplifier and new guarding and shielding techniques were used in the picoammeter circuit which allowed as to measure current less than 1 pA with a current gain of 0.22 V/pA and noise less than 10 fA. This paper will also describe a novel compensation approach which reduced the charge injection from switches in the GI to 10 −18 C, and a T-switch configuration which was used to eliminate leakage current in the reset switch.

  2. A Device for Automatically Measuring and Supervising the Critical Care Patient’S Urine Output

    Directory of Open Access Journals (Sweden)

    Roemi Fernández

    2010-01-01

    Full Text Available Critical care units are equipped with commercial monitoring devices capable of sensing patients’ physiological parameters and supervising the achievement of the established therapeutic goals. This avoids human errors in this task and considerably decreases the workload of the healthcare staff. However, at present there still is a very relevant physiological parameter that is measured and supervised manually by the critical care units’ healthcare staff: urine output. This paper presents a patent-pending device capable of automatically recording and supervising the urine output of a critical care patient. A high precision scale is used to measure the weight of a commercial urine meter. On the scale’s pan there is a support frame made up of Bosch profiles that isolates the scale from force transmission from the patient’s bed, and guarantees that the urine flows properly through the urine meter input tube. The scale’s readings are sent to a PC via Bluetooth where an application supervises the achievement of the therapeutic goals. The device is currently undergoing tests at a research unit associated with the University Hospital of Getafe in Spain.

  3. Arrester Resistive Current Measuring System Based on Heterogeneous Network

    Science.gov (United States)

    Zhang, Yun Hua; Li, Zai Lin; Yuan, Feng; Hou Pan, Feng; Guo, Zhan Nan; Han, Yue

    2018-03-01

    Metal Oxide Arrester (MOA) suffers from aging and poor insulation due to long-term impulse voltage and environmental impact, and the value and variation tendency of resistive current can reflect the health conditions of MOA. The common wired MOA detection need to use long cables, which is complicated to operate, and that wireless measurement methods are facing the problems of poor data synchronization and instability. Therefore a novel synchronous measurement system of arrester current resistive based on heterogeneous network is proposed, which simplifies the calculation process and improves synchronization, accuracy and stability and of the measuring system. This system combines LoRa wireless network, high speed wireless personal area network and the process layer communication, and realizes the detection of arrester working condition. Field test data shows that the system has the characteristics of high accuracy, strong anti-interference ability and good synchronization, which plays an important role in ensuring the stable operation of the power grid.

  4. Liner velocity, current, and symmetry measurements on the 32 MA flux compression generator experiment ALT-1

    CERN Document Server

    Clark, D A; Rodríguez, G; Tabaka, L J

    2001-01-01

    Summary form only given, as follows. A flux compression generator based pulse power system, designed, built, and fielded by a Russian team at the All Russian Scientific Research Institute of Experimental Physics (VNIIEF), was used to successfully drive an aluminum liner to velocities greater than 10 km/sec. The experiment objective was to demonstrate performance of a precision liner implosion at Atlas current of 30 MA or greater. Diagnostics to measure liner performance were an essential part of the experiment. An experimental team from Los Alamos National Laboratory (LANL) provided a suite of diagnostics to measure liner performance. Three diagnostics were fielded. 1. a velocity interferometer (VISAR) to continuously measure the liner inner surface velocity from throughout the entire range of travel. 2. Two Faraday rotation devices to measure liner current during the implosion. 3. Sixteen fiber optic impact pins to record liner impact time and provide axial and azimuthal symmetry information. All diagnostics...

  5. Reliability of Sleep Measures from Four Personal Health Monitoring Devices Compared to Research-Based Actigraphy and Polysomnography

    Directory of Open Access Journals (Sweden)

    Janna Mantua

    2016-05-01

    Full Text Available Polysomnography (PSG is the “gold standard” for monitoring sleep. Alternatives to PSG are of interest for clinical, research, and personal use. Wrist-worn actigraph devices have been utilized in research settings for measures of sleep for over two decades. Whether sleep measures from commercially available devices are similarly valid is unknown. We sought to determine the validity of five wearable devices: Basis Health Tracker, Misfit Shine, Fitbit Flex, Withings Pulse O2, and a research-based actigraph, Actiwatch Spectrum. We used Wilcoxon Signed Rank tests to assess differences between devices relative to PSG and correlational analysis to assess the strength of the relationship. Data loss was greatest for Fitbit and Misfit. For all devices, we found no difference and strong correlation of total sleep time with PSG. Sleep efficiency differed from PSG for Withings, Misfit, Fitbit, and Basis, while Actiwatch mean values did not differ from that of PSG. Only mean values of sleep efficiency (time asleep/time in bed from Actiwatch correlated with PSG, yet this correlation was weak. Light sleep time differed from PSG (nREM1 + nREM2 for all devices. Measures of Deep sleep time did not differ from PSG (SWS + REM for Basis. These results reveal the current strengths and limitations in sleep estimates produced by personal health monitoring devices and point to a need for future development.

  6. Design and Functional Validation of a Complex Impedance Measurement Device for Characterization of Ultrasonic Transducers

    International Nuclear Information System (INIS)

    De-Cock, Wouter; Cools, Jan; Leroux, Paul

    2013-06-01

    This paper presents the design and practical implementation of a complex impedance measurement device capable of characterization of ultrasonic transducers. The device works in the frequency range used by industrial ultrasonic transducers which is below the measurement range of modern high end network analyzers. The device uses the Goertzel algorithm instead of the more common FFT algorithm to calculate the magnitude and phase component of the impedance under test. A theoretical overview is given followed by a practical approach and measurement results. (authors)

  7. CURRENT STATE ANALYSIS OF AUTOMATIC BLOCK SYSTEM DEVICES, METHODS OF ITS SERVICE AND MONITORING

    Directory of Open Access Journals (Sweden)

    A. M. Beznarytnyy

    2014-01-01

    Full Text Available Purpose. Development of formalized description of automatic block system of numerical code based on the analysis of characteristic failures of automatic block system and procedure of its maintenance. Methodology. For this research a theoretical and analytical methods have been used. Findings. Typical failures of the automatic block systems were analyzed, as well as basic reasons of failure occur were found out. It was determined that majority of failures occurs due to defects of the maintenance system. Advantages and disadvantages of the current service technology of automatic block system were analyzed. Works that can be automatized by means of technical diagnostics were found out. Formal description of the numerical code of automatic block system as a graph in the state space of the system was carried out. Originality. The state graph of the numerical code of automatic block system that takes into account gradual transition from the serviceable condition to the loss of efficiency was offered. That allows selecting diagnostic information according to attributes and increasing the effectiveness of recovery operations in the case of a malfunction. Practical value. The obtained results of analysis and proposed the state graph can be used as the basis for the development of new means of diagnosing devices for automatic block system, which in turn will improve the efficiency and service of automatic block system devices in general.

  8. Analysis of insertion device magnet measurements for the Advanced Light Source

    International Nuclear Information System (INIS)

    Marks, S.; Humphries, D.; Kincaid, B.M.; Schlueter, R.; Wang, C.

    1993-07-01

    The Advanced Light Source (ALS), which is currently being commissioned at Lawrence Berkeley Laboratory, is a third generation light source designed to produce XUV radiation of unprecedented brightness. To meet the high brightness goal the storage ring has been designed for very small electron beam emittance and the undulators installed in the ALS are built to a high degree of precision. The allowable magnetic field errors are driven by electron beam and radiation requirements. Detailed magnetic measurements and adjustments are performed on each undulator to qualify it for installation in the ALS. The first two ALS undulators, IDA and IDB, have been installed. This paper describes the program of measurements, data analysis, and adjustments carried out for these two devices. Calculations of the radiation spectrum, based upon magnetic measurements, are included. Final field integral distributions are also shown. Good field integral uniformity has been achieved using a novel correction scheme, which is also described

  9. Four-point probe measurements using current probes with voltage feedback to measure electric potentials

    Science.gov (United States)

    Lüpke, Felix; Cuma, David; Korte, Stefan; Cherepanov, Vasily; Voigtländer, Bert

    2018-02-01

    We present a four-point probe resistance measurement technique which uses four equivalent current measuring units, resulting in minimal hardware requirements and corresponding sources of noise. Local sample potentials are measured by a software feedback loop which adjusts the corresponding tip voltage such that no current flows to the sample. The resulting tip voltage is then equivalent to the sample potential at the tip position. We implement this measurement method into a multi-tip scanning tunneling microscope setup such that potentials can also be measured in tunneling contact, allowing in principle truly non-invasive four-probe measurements. The resulting measurement capabilities are demonstrated for \

  10. Time structure measurement of the ATLAS RPC gap current

    CERN Document Server

    Aielli, G; The ATLAS collaboration

    2010-01-01

    The current absorbed by an RPC represents the sum of the charge delivered in the gas by the ionizing events interesting the gap, integrated by the electrodes time constant. This is typically of the order of tens of ms thus dominating the gas discharge time scale and characterizing the granular structure observed in the current signal. In most cases this structure is considered as noise to be further integrated to observe the average gap current, used often as a detector monitoring parameter or to precisely measure the uncorrelated background rate effects. A remarkable case is given if a large number of particles is passing trough the detector within an integration time constant producing a current peak clearly detectable above the average noise. The ATLAS RPC system is equipped with a dedicated current monitoring based on an ADC capable of reading out the average value as well as the transient peaks of the currents above a given threshold. A study on such data was used to spot the gap HV noise, to monitor the...

  11. Electronic alarm device for radioactivity detector associated with a direct current amplifier or with a integration-based counting assembly

    International Nuclear Information System (INIS)

    Desmaretz, Marc; Ferlicot, Rene

    1964-04-01

    The authors report the study of a device aimed at triggering sound and light alarms when a radiation detector associated with a direct current amplifier or with a counting assembly detects a radiation intensity greater than one or two previously defined thresholds. This device can be used at any time for a detection assembly which is not continuously monitored. It has been designed to be adapted to the CEA standard electronics currently used in installations and on which the alarm function had not been initially foreseen. The assembly comprises an additional safety device for the control of any untimely shutdown of the detection chain [fr

  12. Analysis of operations and cyber security policies for a system of cooperating Flexible Alternating Current Transmission System (FACTS) devices.

    Energy Technology Data Exchange (ETDEWEB)

    Phillips, Laurence R.; Tejani, Bankim; Margulies, Jonathan; Hills, Jason L.; Richardson, Bryan T.; Baca, Micheal J.; Weiland, Laura

    2005-12-01

    Flexible Alternating Current Transmission Systems (FACTS) devices are installed on electric power transmission lines to stabilize and regulate power flow. Power lines protected by FACTS devices can increase power flow and better respond to contingencies. The University of Missouri Rolla (UMR) is currently working on a multi-year project to examine the potential use of multiple FACTS devices distributed over a large power system region in a cooperative arrangement in which the FACTS devices work together to optimize and stabilize the regional power system. The report describes operational and security challenges that need to be addressed to employ FACTS devices in this way and recommends references, processes, technologies, and policies to address these challenges.

  13. Variations in the electrical short-circuit current decay for recombination lifetime and velocity measurements

    Science.gov (United States)

    Jung, Tae-Won; Lindholm, Fredrik A.; Neugroschel, Arnost

    1987-01-01

    An improved measurement system for electrical short-circuit current decay is presented that extends applicability of the method to silicon solar cells having an effective lifetime as low as 1 microsec. The system uses metal/oxide/semiconductor transistors as voltage-controlled switches. Advances in theory developed here increase precision and sensitivity in the determination of the minority-carrier recombination lifetime and recombination velocity. A variation of the method, which exploits measurements made on related back-surface field and back-ohmic contact devices, further improves precision and sensitivity. The improvements are illustrated by application to 15 different silicon solar cells.

  14. A simple indentation device for measuring micrometer-scale tissue stiffness

    Energy Technology Data Exchange (ETDEWEB)

    Levental, I; Levental, K R; Janmey, P A [Institute for Medicine and Engineering, University of Pennsylvania, Philadelphia, PA 19104 (United States); Klein, E A; Assoian, R [Department of Pharmacology, University of Pennsylvania, Philadelphia, PA 19104 (United States); Miller, R T [Departments of Medicine and Physiology, Louis Stokes VAMC, Cleveland, OH (United States); Wells, R G, E-mail: janmey@mail.med.upenn.ed [Department of Medicine, University of Pennsylvania, Philadelphia, PA 19104 (United States)

    2010-05-19

    Mechanical properties of cells and extracellular matrices are critical determinants of function in contexts including oncogenic transformation, neuronal synapse formation, hepatic fibrosis and stem cell differentiation. The size and heterogeneity of biological specimens and the importance of measuring their mechanical properties under conditions that resemble their environments in vivo present a challenge for quantitative measurement. Centimeter-scale tissue samples can be measured by commercial instruments, whereas properties at the subcellular (nm) scale are accessible by atomic force microscopy, optical trapping, or magnetic bead microrheometry; however many tissues are heterogeneous on a length scale between micrometers and millimeters which is not accessible to most current instrumentation. The device described here combines two commercially available technologies, a micronewton resolution force probe and a micromanipulator for probing soft biological samples at sub-millimeter spatial resolution. Several applications of the device are described. These include the first measurement of the stiffness of an intact, isolated mouse glomerulus, quantification of the inner wall stiffness of healthy and diseased mouse aortas, and evaluation of the lateral heterogeneity in the stiffness of mouse mammary glands and rat livers with correlation of this heterogeneity with malignant or fibrotic pathology as evaluated by histology.

  15. Performance of current-in-plane pseudo-spin-valve devices on CMOS silicon-on-insulator underlayers

    Science.gov (United States)

    Katti, R. R.; Zou, D.; Reed, D.; Schipper, D.; Hynes, O.; Shaw, G.; Kaakani, H.

    2003-05-01

    Prior work has shown that current-in-plane (CIP) giant magnetoresistive (GMR) pseudo-spin-valve (PSV) devices grown on bulk Si wafers and bulk complementary metal-oxide semiconductor (CMOS) underlayers exhibit write and read characteristics that are suitable for application as nonvolatile memory devices. In this work, CIP GMR PSV devices fabricated on silicon-on-insulator CMOS underlayers are shown to support write and read performance. Reading and writing fields for selected devices are shown to be approximately 25%-50% that of unselected devices, which provides a margin for reading and writing specific bits in a memory without overwriting bits and without disturbing other bits. The switching characteristics of experimental devices were compared to and found to be similar with Landau-Lifschitz-Gilbert micromagnetic modeling results, which allowed inferring regions of reversible and irreversible rotations in magnetic reversal processes.

  16. Resistivity measurements using a direct current induction method (1963)

    International Nuclear Information System (INIS)

    Delaplace, J.; Hillairet, J.

    1964-01-01

    The conventional methods for measuring electrical resistivities necessitate the fixing of electrical contacts on the sample either mechanically or by soldering. Furthermore it is also necessary to carry,out the measurements on low cross-section samples which are not always easy to obtain. Our direct-current induction method on the other hand requires no contacts and can easily be applied to samples of large cross-section. The sample is placed in a uniform magnetic field; at the moment when the current is cut, eddy currents appear in the sample which tend to oppose the disappearance of the field. The way in which the magnetic flux decreases in the sample makes it possible to determine the resistivity of the material. This method has been applied to samples having diameters of between 1 and 30 mm in the case of metals which are good conductors. It gives a value for the local resistivity and makes it possible to detect any variation along a sample. The measurements can be carried out at all temperature from a few degrees absolute to 500 deg. C. We have used the induction method to follow the purification of beryllium by zone-melting; it is in effect possible to estimate the purity of a material by resistivity measurements. We have measured the resistivity along each bar treated by the zone-melting technique and have thus, localised the purest section. High temperature measurements have been carried out on uranium carbide and on iron-aluminium alloys. This method constitutes an interesting means of investigation the resistivity of solid materials. Its accuracy and rapidity make it particularly adapted both to fundamental research and to production control. (authors) [fr

  17. Design, Assembly, and Commissioning of a Cryogenic DC Current Transformer Designed for Measuring Currents of up to 80 kA

    CERN Document Server

    Montenero, G; Bottura, L; Arpaia, P

    2015-01-01

    A new cryogenic dc current transformer (Cryo-DCCT) has recently been designed and assembled at CERN. The device, whose design is based on that of a high-accuracy 600 A market solution suitable for room temperature applications, is optimized for measuring currents of up to 80 kA and for operation at 4.2 K. The CryoDCCT has been conceived with the objective of preserving the metrological performance of the original commercial device in the new extended range of operation. For reducing the effect of interfering magnetic fields arising from test conditions, it incorporates ferromagnetic and MgB2 superconducting shields. In this paper, the design of the CryoDCCT and the results of the commissioning of the device at CERN are reported. The effectiveness of the current transducer is analysed and discussed. This new device will be used for measuring the secondary current of a 80 kA superconducting transformer feeding a sample of NbSn3 cable at the Facility for Research on Superconducting Cables (FRESCA) at CERN.

  18. Current state of micro-robots/devices as substitutes for screening colonoscopy: assessment based on technology readiness levels.

    Science.gov (United States)

    Tapia-Siles, Silvia C; Coleman, Stuart; Cuschieri, Alfred

    2016-02-01

    Previous reports have described several candidates, which have the potential to replace colonoscopy, but to date, there is still no device capable of fully replacing flexible colonoscopy in the management of colonic disorders and for mass adult population screening for asymptomatic colorectal cancer. NASA developed the TRL methodology to describe and define the stages of development before use and marketing of any device. The definitions of the TRLS used in the present review are those formulated by "The US Department of Defense Technology Readiness Assessment Guidance" but adapted to micro-robots for colonoscopy. All the devices included are reported in scientific literature. They were identified by a systematic search in Web of Science, PubMed and IEEE Xplore amongst other sources. Devices that clearly lack the potential for full replacement of flexible colonoscopy were excluded. The technological salient features of all the devices included for assessment are described briefly, with particular focus on device propulsion. The devices are classified according to the TRL criteria based on the reported information. An analysis is next undertaken of the characteristics and salient features of the devices included in the review: wireless/tethered devices, data storage-transmission and navigation, additional functionality, residual technology challenges and clinical and socio-economical needs. Few devices currently possess the required functionality and performance to replace the conventional colonoscopy. The requirements, including functionalities which favour the development of a micro-robot platform to replace colonoscopy, are highlighted.

  19. Development of a prototype solid state fault current limiting and interrupting device for low voltage distribution networks.

    OpenAIRE

    Ahmed, M.; Putrus, G. A.; Ran, L.; Penlington, R.

    2006-01-01

    This paper describes the development of a solid-state Fault Current Limiting and Interrupting Device (FCLID) suitable for low voltage distribution networks. The main components of the FCLID are a bidirectional semiconductor switch that can disrupt the short-circuit current, and a voltage clamping element that helps in controlling the current and absorbing the inductive energy stored in the network during current interruption. Using a hysteresis type control algorithm, the short-circuit curren...

  20. Quantifying short-lived events in multistate ionic current measurements.

    Science.gov (United States)

    Balijepalli, Arvind; Ettedgui, Jessica; Cornio, Andrew T; Robertson, Joseph W F; Cheung, Kin P; Kasianowicz, John J; Vaz, Canute

    2014-02-25

    We developed a generalized technique to characterize polymer-nanopore interactions via single channel ionic current measurements. Physical interactions between analytes, such as DNA, proteins, or synthetic polymers, and a nanopore cause multiple discrete states in the current. We modeled the transitions of the current to individual states with an equivalent electrical circuit, which allowed us to describe the system response. This enabled the estimation of short-lived states that are presently not characterized by existing analysis techniques. Our approach considerably improves the range and resolution of single-molecule characterization with nanopores. For example, we characterized the residence times of synthetic polymers that are three times shorter than those estimated with existing algorithms. Because the molecule's residence time follows an exponential distribution, we recover nearly 20-fold more events per unit time that can be used for analysis. Furthermore, the measurement range was extended from 11 monomers to as few as 8. Finally, we applied this technique to recover a known sequence of single-stranded DNA from previously published ion channel recordings, identifying discrete current states with subpicoampere resolution.

  1. AN INDUCTION SENSOR FOR MEASURING CURRENTS OF NANOSECOND RANGE

    Directory of Open Access Journals (Sweden)

    S. P. Shalamov

    2016-11-01

    Full Text Available Purpose. A current meter based on the principle of electromagnetic induction is designed to register the current flowing in the rod lightning. The aim of the article is to describe the way of increasing the sensitivity of the converter by means of their serial communication. Methodology. The recorded current is in the nanosecond range. If compared with other methods, meters based on the principle of electromagnetic induction have several advantages, such as simplicity of construction, reliability, low cost, no need in a power source, relatively high sensitivity. Creation of such a meter is necessary, because in some cases there is no possibility to use a shunt. Transient properties of a meter are determined by the number of turns and the constant of integration. Sensitivity is determined by measuring the number of turns, the coil sectional area, the core material and the integration constant. For measuring the magnetic field pulses with a rise time of 5 ns to 50 ns a meter has turns from 5 to 15. The sensitivity of such a meter is low. When the number of turns is increased, the output signal and the front increase. Earlier described dependencies were used to select the main parameters of the converter. It was based on generally accepted and widely known equivalent circuit. The experience of created earlier pulse magnetic field meters was considered both for measuring the magnetic fields, and large pulse current. Originality. Series connection of converters has the property of a long line. The level of the transient response of the meter is calculated. The influence of parasitic parameters on the type of meter transient response is examined. The shown construction was not previously described. Practical value. The results of meter implementation are given. The design peculiarities of the given measuring instruments are shown.

  2. Clinical decision support systems in hospital care using ubiquitous devices: Current issues and challenges.

    Science.gov (United States)

    Baig, Mirza Mansoor; GholamHosseini, Hamid; Moqeem, Aasia A; Mirza, Farhaan; Lindén, Maria

    2017-11-01

    Supporting clinicians in decision making using advanced technologies has been an active research area in biomedical engineering during the past years. Among a wide range of ubiquitous systems, smartphone applications have been increasingly developed in healthcare settings to help clinicians as well as patients. Today, many smartphone applications, from basic data analysis to advanced patient monitoring, are available to clinicians and patients. Such applications are now increasingly integrating into healthcare for clinical decision support, and therefore, concerns around accuracy, stability, and dependency of these applications are rising. In addition, lack of attention to the clinicians' acceptability, as well as the low impact on the medical professionals' decision making, are posing more serious issues on the acceptability of smartphone applications. This article reviews smartphone-based decision support applications, focusing on hospital care settings and their overall impact of these applications on the wider clinical workflow. Additionally, key challenges and barriers of the current ubiquitous device-based healthcare applications are identified. Finally, this article addresses current challenges, future directions, and the adoption of mobile healthcare applications.

  3. Leakage current measurements on pixelated CdZnTe detectors

    International Nuclear Information System (INIS)

    Dirks, B.P.F.; Blondel, C.; Daly, F.; Gevin, O.; Limousin, O.; Lugiez, F.

    2006-01-01

    In the field of the R and D of a new generation hard X-ray cameras for space applications we focus on the use of pixelated CdTe or CdZnTe semiconductor detectors. They are covered with 64 (0.9x0.9 mm 2 ) or 256 (0.5x0.5 mm 2 ) pixels, surrounded by a guard ring and operate in the energy ranging from several keV to 1 MeV, at temperatures between -20 and +20 o C. A critical parameter in the characterisation of these detectors is the leakage current per pixel under polarisation (∼50-500 V/mm). In operation mode each pixel will be read-out by an integrated spectroscopy channel of the multi-channel IDeF-X ASIC currently developed in our lab. The design and functionality of the ASIC depends directly on the direction and value of the current. A dedicated and highly insulating electronics circuit is designed to automatically measure the current in each individual pixel, which is in the order of tens of pico-amperes. Leakage current maps of different CdZnTe detectors of 2 and 6 mm thick and at various temperatures are presented and discussed. Defect density diagnostics have been performed by calculation of the activation energy of the material

  4. Design and test of 4πβ-γ coincidence measurement device based on DSP technology

    International Nuclear Information System (INIS)

    Zeng Herong; Feng Qijie; Leng Jun; Qian Dazhi; Bai Lixin; Zhang Yiyun

    2012-01-01

    The paper illustrates the hardware and software of the 4πβ-γ coincidence measurement device based on DSP technology in detail. In such device, the single-channel analyzer, gate generator, coincidence circuit and scalar in the traditional coincidence measurement device are replaced by the digital coincidence acquirer which is researched and manufactured by ourselves. Doing so, the measurement efficiency will be respectively improved, and the hardware cost will be lowered. The comparison experiment shows that the design of such device is a success. (authors)

  5. Damping constant measurement and inverse giant magnetoresistance in spintronic devices with Fe4N

    Directory of Open Access Journals (Sweden)

    Xuan Li

    2017-12-01

    Full Text Available Fe4N is one of the attractive materials for spintronic devices due to its large spin asymmetric conductance and negative spin polarization at the Fermi level. We have successfully deposited Fe4N thin film with (001 out-of-plane orientation using a DC facing-target-sputtering system. A Fe(001/Ag(001 composite buffer layer is selected to improve the (001 orientation of the Fe4N thin film. The N2 partial pressure during sputtering is optimized to promote the formation of Fe4N phase. Moreover, we have measured the ferromagnetic resonance (FMR of the (001 oriented Fe4N thin film using coplanar waveguides and microwave excitation. The resonant fields are tested under different microwave excitation frequencies, and the experimental results match well with the Kittel formula. The Gilbert damping constant of Fe4N is determined to be α = 0.021±0.02. We have also fabricated and characterized the current-perpendicular-to-plane (CPP giant magnetoresistance (GMR device with Fe4N/Ag/Fe sandwich. Inverse giant magnetoresistance is observed in the CPP GMR device, which suggests that the spin polarization of Fe4N and Fe4N/Ag interface is negative.

  6. Direct-current proton-beam measurements at Los Alamos

    International Nuclear Information System (INIS)

    Sherman, J.; Stevens, R.R.; Schneider, J.D.; Zaugg, T.

    1994-01-01

    Recently, a CW proton accelerator complex was moved from Chalk River Laboratories (CRL) to Los Alamos National Laboratory. This includes a 50-keV dc proton injector with a single-solenoid low-energy beam transport system (LEBT) and a CW 1.25-MeV, 267-MHz radiofrequency quadrupole (RFQ). The move was completed after CRL had achieved 55-mA CW operation at 1.25 MeV using 250-kW klystrode tubes to power the RFQ. These accelerator components are prototypes for the front end of a CW linac required for an accelerator-driven transmutation linac, and they provide early confirmation of some CW accelerator components. The injector (ion source and LEBT) and emittance measuring unit are installed and operational at Los Alamos. The dc microwave ion source has been operated routinely at 50-keV, 75-mA hydrogen-ion current. This ion source has demonstrated very good discharge and H 2 gas efficiencies, and sufficient reliability to complete CW RFQ measurements at CRL. Proton fraction of 75% has been measured with 550-W discharge power. This high proton fraction removes the need for an analyzing magnet. Proton LEBT emittance measurements completed at Los Alamos suggest that improved transmission through the RFQ may be achieved by increasing the solenoid focusing current. Status of the final CW RFQ operation at CRL and the installation of the RFQ at Los Alamos is given

  7. Measuring current emission and work functions of large thermionic cathodes

    International Nuclear Information System (INIS)

    Fortgang, C.M.

    2001-01-01

    As one component of the nations Stockpile Stewardship program, Los Alamos National Laboratory is constructing a 20 MeV, 2 kA (with a 4 kA upgrade capability), 3ps induction linac for doing x-ray radiography of explosive devices. The linac is one leg of a facility called the Dual-Axis Radiography Hydrodynamic Test Facility (DARHT). The electron gun is designed to operate at 3.2 MV. The gun is a Pierce type design and uses a 6.5' cathode for 2 kA operation and an 8' cathode for 4 kA operation. We have constructed a small facility called the Cathode Test Stand (CTS) to investigate engineering and physics issues regarding large thermionic dispenser-cathodes. In particular, we have looked at the issues of temperature uniformity on the cathode surface and cathode quality as measured by its work function. We have done thermal imaging of both 8' and 6.5' cathodes. Here we report on measurements of the cathode work function, both the average value and how it vanes across the face of the cathode.

  8. Digital Measuring Devices Used for Injector Hydraulic Test

    Directory of Open Access Journals (Sweden)

    S. N. Leontiev

    2015-01-01

    Full Text Available To ensure a high specific impulse of the LRE (liquid-propellant engine chamber it is necessary to have optimally organized combustion of the fuel components. This can be ensured by choosing the optimum geometry of gas-dynamic contour of the LRE combustor, as well as by improving the sputtering processes and mixing the fuel components, for example, by selection of the optimum type, characteristics, and location of injectors on the mixing unit of the chamber.These particular reasons arise the interest in the injector characteristics in terms of science, and technological aspects determine the need for control of underlying design parameters in their manufacture.The objective of this work is to give an experimental justification on used digital measurement instrumentation and research the hydraulic characteristics of injectors.To determine injector parameters most widely were used the units with sectional collectors. A technique to control injector parameters using the sectional collectors involves spraying the liquid by injector at a given pressure drop on it for a certain time (the longer, the higher the accuracy and reliability of the results and then determining the amount of liquid in each section to calculate the required parameters of injector.In this work the liquid flow through the injector was determined by high-precision flowmeters FLONET FN2024.1 of electromagnetic type, which have very high metrological characteristics, in particular a flow rate error does not exceed 0.5% in a range of water flow from Qmin= 0.0028 l/s to Qmax Qmax = 0.28 l/s. To determine the coefficient of uneven spray were used differential pressure sensors DMD 331-ASLX of company "DB Sensors RUS", which have an error of 0.075% with a range of differential pressure 0 ... 5 kPa. Measuring complex MIC-200 of company "NPP Measure" and WinPos software for processing array information provided entry, recording, and processing of all the data of the experiment.In this

  9. Assessment of a portable device for the quantitative measurement of ankle joint stiffness in spastic individuals

    DEFF Research Database (Denmark)

    Lorentzen, Jakob; Grey, Michael J; Geertsen, Svend Sparre

    2012-01-01

    was measured with the portable device and a stationary torque motor. Inter- and intra-rater reliability was assessed with the intra-class correlation coefficient (ICC). RESULTS: Stiffness measures with the portable and stationary devices were significantly correlated for controls and MS participants (p...

  10. Measuring devices for the modular switch system; Messgeraete fuer den Schaltschrank

    Energy Technology Data Exchange (ETDEWEB)

    Mueller, Rudolf [Janitza Electronics GmbH, Lahnau (Germany). Sales und Marketing

    2008-10-15

    The advantages of digital universal measuring instruments are: lower device cost for more information and functionality. Furtheron digital measuring technology is more exactly during service life. Cost advantages result due to low installation cost and reduced installation of wires and cables. So universal devices replace all analogue systems and offer further functions. (orig./GL)

  11. 30 CFR 75.320 - Air quality detectors and measurement devices.

    Science.gov (United States)

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Air quality detectors and measurement devices... COAL MINE SAFETY AND HEALTH MANDATORY SAFETY STANDARDS-UNDERGROUND COAL MINES Ventilation § 75.320 Air quality detectors and measurement devices. (a) Tests for methane shall be made by a qualified person with...

  12. Recommendations for blood pressure measuring devices for office/clinic use in low resource settings

    NARCIS (Netherlands)

    Parati, G.; Mendis, S.; Abegunde, D.; Asmar, R.; Mieke, S.; Murray, A.; Shengelia, B.; Steenvoorden, G.; Montfrans, G. van; O'Brien, E.

    2005-01-01

    This paper, which summarizes the conclusions of a WHO Expert meeting, is aimed at proposing indications to develop technical specifications for an accurate and affordable blood pressure measuring device for office/ clinic use in low resource settings. Blood pressure measuring devices to be used in

  13. Comparisons of switching characteristics between Ti/Al2O3/Pt and TiN/Al2O3/Pt RRAM devices with various compliance currents

    Science.gov (United States)

    Qi, Yanfei; Zhao, Ce Zhou; Liu, Chenguang; Fang, Yuxiao; He, Jiahuan; Luo, Tian; Yang, Li; Zhao, Chun

    2018-04-01

    In this study, the influence of the Ti and TiN top electrodes on the switching behaviors of the Al2O3/Pt resistive random access memory devices with various compliance currents (CCs, 1-15 mA) has been compared. Based on the similar statistical results of the resistive switching (RS) parameters such as V set/V reset, R HRS/R LRS (measured at 0.10 V) and resistance ratio with various CCs for both devices, the Ti/Al2O3/Pt device differs from the TiN/Al2O3/Pt device mainly in the forming process rather than in the following switching cycles. Apart from the initial isolated state, the Ti/Al2O3/Pt device has the initial intermediate state as well. In addition, its forming voltage is relatively lower. The conduction mechanisms of the ON and OFF state for both devices are demonstrated as ohmic conduction and Frenkel-Poole emission, respectively. Therefore, with the combined modulations of the CCs and the stop voltages, the TiN/Al2O3/Pt device is more stable for nonvolatile memory applications to further improve the RS performance.

  14. Experimental device for measuring the dynamic properties of diaphragm motors

    Science.gov (United States)

    Fojtášek, Kamil; Dvořák, Lukáš; Mejzlík, Jan

    The subject of this paper is to design and description of the experimental device for the determination dynamic properties of diaphragm pneumatic motors. These motors are structurally quite different from conventional pneumatic linear cylinders. The working fluid is typically compressed air, the piston of motor is replaced by an elastic part and during the working cycle there is a contact of two elastic environments. In the manufacturers catalogs of these motors are not given any working characteristics. Description of the dynamic behavior of diaphragm motor will be used for verification of mathematical models.

  15. Accuracy of an improved device for remote measuring of tree-trunk diameters

    International Nuclear Information System (INIS)

    Matsushita, T.; Kato, S.; Komiyama, A.

    2000-01-01

    For measuring the diameters of tree trunks from a distant position, a recent device using a laser beam was developed by Kantou. We improved this device to serve our own practical purposes. The improved device consists of a 1-m-long metal caliper and a small telescope sliding smoothly onto it. Using the cross hairs in the scope, one can measure both edges of an object on the caliper and calculate its length. The laser beam is used just for guiding the telescopic sights to the correct positions on the object. In this study, the accuracy of this new device was examined by measuring objects of differing lengths, the distance from the object, and the angle of elevation to the object. Since each result of the experiment predicted absolute errors of measurement of less than 3 mm, this new device will be suitable for the measurement of trunk diameters in the field

  16. Poloidal polarimeter for current density measurements in ITER

    International Nuclear Information System (INIS)

    Donne, A.J.H.; Graswinckel, M.F.; Cavinato, M.; Giudicotti, L.; Zilli, E.; Gil, C.; Koslowski, H.R.; McCarthy, P.; Nyhan, C.; Prunty, S.; Spillane, M.; Walker, C.

    2004-01-01

    One of the systems envisaged for measuring the current density profile in the ITER is a 118 μm poloidal polarimeter system. The proposed system has two independent views: one fan of chords observes the plasma via an equatorial port and a second fan views down from an upper port. This article will present the status of the on-going work and will address issues as sensitivity and accuracy, refraction, Gaussian beam ray-tracing, alignment, and calibration as well as some specific design details

  17. Development of a Force Measurement Device for Lower-Body Muscular Strength Measuring of Skaters

    Science.gov (United States)

    Kim, Dong Ki; Lee, Jeong Tae

    This paper presents a force measurement system that can measure a lower-body muscular strength of skaters. The precise measurement and analysis of the left and right lower-body strength of skaters is necessary, because a left/right lower-body strength balance is helpful to improve the athletes' performance and to protect them from injury. The system is constructed with a skate sliding board, a couple of sensor-units with load cell, indicator and control box, guard, force pad, and support bracket. The developed force measurement system is calibrated by the calibration setup, and the uncertainty of the force sensing unit on the left is within 0.087% and the uncertainty of the force sensing unit on the right is within 0.109%. In order to check the feasibility of the developed measurement device, a kinematic analysis is conducted with skater. As a result, the subject shows the deviation of left and right of 12.1 N with respect to average strength and 39.1 N with respect to the maximum strength. This evaluation results are reliable enough to make it possible to measure a lower-body muscular strength of skaters. The use of this measurement system will be expected to correct the posture of skaters and record the sports dynamics data for each athlete. It is believed that through the development of this equipment, skaters in elementary, middle, high schools, colleges, and the professional level have the systematic training to compete with world-class skaters.

  18. Device for measuring the flow rate of a fluid moving through a pipe

    International Nuclear Information System (INIS)

    Barge, Gilles; Bouchard, Patrick; Chaix, J.E.; Rigaud, J.L.; Vivaldi, Andre.

    1981-01-01

    A device is described for measuring the flow rate, in particular through large section pipes, such as those found in water type nuclear reactors, thermal power stations and gas loops. This device includes a plate drilled with holes crossed by a fluid and held in the pipe by deformable components on which are secured strain gauges forming the detecting element of an electronic device for processing the signal emitted by the gauges. This device can be employed, for instance, for measuring the flow rate of a coolant in the primary system of a nuclear reactor [fr

  19. Development and evaluation of measurement devices used to support testing of radioactive material transportation packages

    International Nuclear Information System (INIS)

    Uncapher, W.L.; Ammerman, D.J.; Stenberg, D.R.; Bronowski, D.R.; Arviso, M.

    1992-01-01

    Radioactive material package designers use structural testing to verify and demonstrate package performance. A major part of evaluating structural response is the collection of instrumentation measurement data. Sandia National Laboratories (SNL) has an ongoing program to develop and evaluate measurement devices to support testing of radioactive material packages. Measurement devices developed in support of this activity include evaluation channels, ruggedly constructed linear variable differential transformers, and piezoresistive accelerometers with enhanced measurement capabilities. In addition to developing measurement devices, a method has been derived to evaluate accelerometers and strain gages for measurement repeatability, ruggedness, and manufacturers' calibration data under both laboratory and field conditions. The developed measurement devices and evaluation technique will be discussed and the results of the evaluation will be presented

  20. Simultaneous specimen current and time-dependent cathodoluminescence measurements on gallium nitride

    Energy Technology Data Exchange (ETDEWEB)

    Campo, E. M., E-mail: e.campo@bangor.ac.uk; Hopkins, L. [School of Electronic Engineering, Bangor University, Gwynedd LL57 1UT (United Kingdom); Pophristic, M. [Department of Chemistry and Biochemistry, University of the Science, Philadelphia, Pennsylvania 19104 (United States); Ferguson, I. T. [Department of Electrical and Computer Engineering, Missouri University of Science and Technology, Rolla, Missouri 65409 (United States)

    2016-06-28

    Time-dependent cathodoluminescence (CL) and specimen current (SC) are monitored to evaluate trapping behavior and evolution of charge storage. Examination of CL and SC suggests that the near band edge emission in GaN is reduced primarily by the activation of traps upon irradiation, and Gallium vacancies are prime candidates. At the steady state, measurement of the stored charge by empiric-analytical methods suggests that all available traps within the interaction volume have been filled, and that additional charge is being stored interstitially, necessarily beyond the interaction volume. Once established, the space charge region is responsible for the steady state CL emission and, prior to build up, it is responsible for the generation of diffusion currents. Since the non-recombination effects resulting from diffusion currents that develop early on are analogous to those leading to device failure upon aging, this study is fundamental toward a holistic insight into optical properties in GaN.

  1. Development of system for automatic measurement of transient photocurrent and thermally stimulated current

    Directory of Open Access Journals (Sweden)

    Asdrubal Antonio Ramirez Botero

    2017-01-01

    Full Text Available This paper presents details of the design and implementation of a system for measuring of thermally stimulated current (TSC and transient photocurrent (Iph, developed using the Virtual Instrumentation concept. For that we have used National Instrument hardware and the LabView® package as software. The system is controlled by a virtual instrument (VI which includes facilities to perform measurements of photocurrent keeping the temperature of the sample and the pressure of the chamber of measurement controlled as well as  real time display of the Iph vs t and TSC vs T curves. The system was tested by performing transient photocurrent and TSC measurements on CH3NH3PbI3 thin films that are generally used as absorbent layer of solar cells. This type of characterization is very useful to get information of the  trapping and recombination processes that affect the transport properties of the devices.

  2. Validity and Reliability of a New Device (WIMU®) for Measuring Hamstring Muscle Extensibility.

    Science.gov (United States)

    Muyor, José M

    2017-09-01

    The aims of the current study were 1) to evaluate the validity of the WIMU ® system for measuring hamstring muscle extensibility in the passive straight leg raise (PSLR) test using an inclinometer for the criterion and 2) to determine the test-retest reliability of the WIMU ® system to measure hamstring muscle extensibility during the PSLR test. 55 subjects were evaluated on 2 separate occasions. Data from a Unilever inclinometer and WIMU ® system were collected simultaneously. Intraclass correlation coefficients (ICCs) for the validity were very high (0.983-1); a very low systematic bias (-0.21°--0.42°), random error (0.05°-0.04°) and standard error of the estimate (0.43°-0.34°) were observed (left-right leg, respectively) between the 2 devices (inclinometer and the WIMU ® system). The R 2 between the devices was 0.999 (p<0.001) in both the left and right legs. The test-retest reliability of the WIMU ® system was excellent, with ICCs ranging from 0.972-0.995, low coefficients of variation (0.01%), and a low standard error of the estimate (0.19-0.31°). The WIMU ® system showed strong concurrent validity and excellent test-retest reliability for the evaluation of hamstring muscle extensibility in the PSLR test. © Georg Thieme Verlag KG Stuttgart · New York.

  3. Development of an Optical Sensor Head for Current and Temperature Measurements in Power Systems

    Directory of Open Access Journals (Sweden)

    Fábio V. B. de Nazaré

    2013-01-01

    Full Text Available The development of a current and temperature monitoring optical device intended to be used in high-voltage environments, particularly transmission lines, is presented. The system is intended to offer not only measurement reliability, but to be also practical and light weighted. Fiber Bragg gratings (FBGs are employed in the measurement of both physical parameters: the current will be acquired using a hybrid sensor head setup—an FBG fixed on a magnetostrictive rod—while a single-point temperature information is provided by a dedicated grating. An inexpensive and outdoor-suitable demodulation method, such as the fixed filter technique, should be used in order to improve the instrumentation robustness, avoiding expensive and complex auxiliary electronics. The preliminary results for laboratory tests are also discussed.

  4. Measurement of electron- and ion beam energies and currents in a plasma focus discharge

    International Nuclear Information System (INIS)

    Yamamoto, Toshikazu; Kondoh, Yoshiomi; Shimoda, Katsuji; Hirano, Katsumi

    1982-01-01

    Measurements of energetic particle beams in a plsma focus with a Mather type device are presented. Rogowski coils are used for time-resolved measurement, and solid-state nuclear track detectors for time-integrated measurement of the beams. In the upstream direction with respect to the discharge current, only the electron beam with the maximum current of several kA was detected, which was approximately one percent of the discharge current. The electron energies of the beam were spread from 0.1 to 1 MeV. In the downstream direction, two successive emissions of ions were observed. The first emission had an extremely high energy of the order of some MeV and a low beam current of less than 10 A. The second emission, the main part of the ion beam, with energies of 100 - 800 keV, followed the first one with a time lag of several tens of nanoseconds, and the beam current reached several tens of amperes. (author)

  5. Field Measurements to Characterize Turbulent Inflow for Marine Hydrokinetic Devices - Marrowstone Island, WA

    Science.gov (United States)

    Richmond, M. C.; Thomson, J. M.; Durgesh, V.; Polagye, B. L.

    2011-12-01

    Field measurements are essential for developing an improved understanding of turbulent inflow conditions that affect the design and operation of marine and hydrokinetic (MHK) devices. The Marrowstone Island site in Puget Sound, Washington State is a potential location for installing MHK devices, as it experiences strong tides and associated currents. Here, field measurements from Nodule Point on the eastern side of Marrowstone Island are used to characterize the turbulence in terms of velocity variance as a function of length and time scales. The field measurements were performed using Acoustic Doppler Velocimetry (ADV) and Acoustic Doppler Current Profiler (ADCP) instruments. Both were deployed on a bottom-mounted tripod at the site by the Applied Physics Lab at the University of Washington (APL-UW). The ADV acquired single point, temporally resolved velocity data from 17-21 Feb 2011, at a height of 4.6 m above the seabed at a sampling frequency of 32 Hz. The ADCP measured the velocity profile over the water column from a height of 2.6 m above the seabed up to the sea-surface in 36 bins, with each bin of 0.5 m size. The ADCP acquired data from 11-27 Feb 2011 at a sampling frequency of 2 Hz. Analysis of the ADV measurements shows distinct dynamic regions by scale: anisotropic eddies at large scales, an isotropic turbulent cascade (-5/3 slope in frequency spectra) at mesoscales, and contamination by Doppler noise at small scales. While Doppler noise is an order of magnitude greater for the ADCP measurements, the turbulence bulk statistics are consistent between the two instruments. There are significant variations in turbulence statistics with stage of the tidal currents (i.e., from slack to non-slack tidal conditions), however an average turbulent intensity of 10% is a robust, canonical value for this site. The ADCP velocity profiles are useful in quantifying the variability in velocity along the water column, and the ensemble averaged velocity profiles may be

  6. Charge Transport in Spiro-OMeTAD Investigated through Space-Charge-Limited Current Measurements

    Science.gov (United States)

    Röhr, Jason A.; Shi, Xingyuan; Haque, Saif A.; Kirchartz, Thomas; Nelson, Jenny

    2018-04-01

    Extracting charge-carrier mobilities for organic semiconductors from space-charge-limited conduction measurements is complicated in practice by nonideal factors such as trapping in defects and injection barriers. Here, we show that by allowing the bandlike charge-carrier mobility, trap characteristics, injection barrier heights, and the shunt resistance to vary in a multiple-trapping drift-diffusion model, a numerical fit can be obtained to the entire current density-voltage curve from experimental space-charge-limited current measurements on both symmetric and asymmetric 2 ,2',7 ,7' -tetrakis(N ,N -di-4-methoxyphenylamine)-9 ,9' -spirobifluorene (spiro-OMeTAD) single-carrier devices. This approach yields a bandlike mobility that is more than an order of magnitude higher than the effective mobility obtained using analytical approximations, such as the Mott-Gurney law and the moving-electrode equation. It is also shown that where these analytical approximations require a temperature-dependent effective mobility to achieve fits, the numerical model can yield a temperature-, electric-field-, and charge-carrier-density-independent mobility. Finally, we present an analytical model describing trap-limited current flow through a semiconductor in a symmetric single-carrier device. We compare the obtained charge-carrier mobility and trap characteristics from this analytical model to the results from the numerical model, showing excellent agreement. This work shows the importance of accounting for traps and injection barriers explicitly when analyzing current density-voltage curves from space-charge-limited current measurements.

  7. Pneumatic strength assessment device: design and isometric measurement.

    Science.gov (United States)

    Paulus, David C; Reiser, Raoul F; Troxell, Wade O

    2004-01-01

    In order to load a muscle optimally during resistance exercise, it should be heavily taxed throughout the entire range of motion for that exercise. However, traditional constant resistance squats only tax the lower-extremity muscles to their limits at the "sticking region" or a critical joint configuration of the exercise cycle. Therefore, a linear motion (Smith) exercise machine was modified with pneumatics and appropriate computer control so that it could be capable of adjusting force to control velocity within a repetition of the squat exercise or other exercise performed with the device. Prior to application of this device in a dynamic squat setting, the maximum voluntary isometric force (MVIF) produced over a spectrum of knee angles is needed. This would reveal the sticking region and overall variation in strength capacity. Five incremental knee angles (90, 110, 130, 150, and 170 degrees, where 180 degrees defined full extension) were examined. After obtaining university-approved informed consent, 12 men and 12 women participated in the study. The knee angle was set, and the pneumatic cylinder was pressurized such that the subject could move the barbell slightly but no more than two-centimeters. The peak pressure exerted over a five-second maximum effort interval was recorded at each knee angle in random order and then repeated. The average of both efforts was then utilized for further analysis. The sticking region occurred consistently at a 90 degrees knee angle, however, the maximum force produced varied between 110 degrees and 170 degrees with the greatest frequency at 150 degrees for both men and women. The percent difference between the maximum and minimum MVIF was 46% for men and 57% for women.

  8. Direct Reconstruction of Two-Dimensional Currents in Thin Films from Magnetic-Field Measurements

    Science.gov (United States)

    Meltzer, Alexander Y.; Levin, Eitan; Zeldov, Eli

    2017-12-01

    An accurate determination of microscopic transport and magnetization currents is of central importance for the study of the electric properties of low-dimensional materials and interfaces, of superconducting thin films, and of electronic devices. Current distribution is usually derived from the measurement of the perpendicular component of the magnetic field above the surface of the sample, followed by numerical inversion of the Biot-Savart law. The inversion is commonly obtained by deriving the current stream function g , which is then differentiated in order to obtain the current distribution. However, this two-step procedure requires filtering at each step and, as a result, oversmooths the solution. To avoid this oversmoothing, we develop a direct procedure for inversion of the magnetic field that avoids use of the stream function. This approach provides enhanced accuracy of current reconstruction over a wide range of noise levels. We further introduce a reflection procedure that allows for the reconstruction of currents that cross the boundaries of the measurement window. The effectiveness of our approach is demonstrated by several numerical examples.

  9. Determination of the energy gap in photoconducting insulators through current noise measurements

    International Nuclear Information System (INIS)

    Carbone, A.; Demichelis, F.; Mazzetti, P.

    1989-01-01

    Measurements of the current noise power spectrum of the photoconducting insulators CdS and CdSe irradiated with monochromatic light of different wavelength λ are reported. It is shown that there is an abrupt change of about one order of magnitude in the low frequency power density of the noise when λ crosses the photoconductor gap value λ and the photocurrent and the device conductance are kept constant by varying the light intensity. The effect is explained in terms of an abrupt increase of the carrier recombination rate when the photon energy becomes larger than the energy gap of the photoconductor. Possible applications of these results are briefly discussed

  10. Non-destructive failure analysis and measurement for molded devices and complex assemblies with X-ray CT and 3D image processing techniques

    International Nuclear Information System (INIS)

    Yin, Xiaoming; Liew, Seaw Jia; Jiang, Ting Ying; Xu, Jian; Kakarala, Ramakrishna

    2013-01-01

    In both automotive and healthcare sectors, reliable failure analysis and accurate measurement of molded devices and complex assemblies are important. Current methods of failure analysis and measurement require these molded parts to be cross-sectioned so that internal features or dimensions can be accessible. As a result, the parts are deemed unusable and additional failure introduced by sectioning may cause misinterpretation of the results. X-ray CT and 3D image processing techniques provide a new nondestructive solution for failure analysis and measurement of molded devices and complex assemblies. These techniques simplify failure analysis and measurement of molded devices and assemblies, and improve the productivity of molding manufacturing significantly.

  11. COMPARISON OF A HEAD MOUNTED IMPACT MEASUREMENT DEVICE TO THE HYBRID III ANTHROPOMORPHIC TESTING DEVICE IN A CONTROLLED LABORATORY SETTING.

    Science.gov (United States)

    Schussler, Eric; Stark, David; Bolte, John H; Kang, Yun Seok; Onate, James A

    2017-08-01

    Reports estimate that 1.6 to 3.8 million cases of concussion occur in sports and recreation each year in the United States. Despite continued efforts to reduce the occurrence of concussion, the rate of diagnosis continues to increase. The mechanisms of concussion are thought to involve linear and rotational head accelerations and velocities. One method of quantifying the kinematics experienced during sport participation is to place measurement devices into the athlete's helmet or directly on the athlete's head. The purpose of this research to determine the accuracy of a head mounted device for measuring the head accelerations experienced by the wearer. This will be accomplished by identifying the error in Peak Linear Acceleration (PLA), Peak Rotational Acceleration (PRA) and Peak Rotational Velocity (PRV) of the device. Laboratory study. A helmeted Hybrid III 50th percentile male headform was impacted via a pneumatic ram from the front, side, rear, front oblique and rear oblique at speeds from 1.5 to 5 m/s. The X2 Biosystems xPatch® (Seattle, WA) sensor was placed on the headform's right side at the approximate location of the mastoid process. Measures of PLA, PRA, PRV from the xPatch ® and Hybrid III were analyzed for Root Mean Square Error (RMSE), and Absolute and Relative Error (AE, RE). Seventy-six impacts were analyzed. All measures of correlation, fixed through the origin, were found to be strong: PLA R 2 =0.967 pstandard yet above the average error of testing devices in both PLA and PRA, but a low error in PRV. PLA measures from the xPatch® system demonstrated a high level of correlation with the PLA data from the Hybrid III mounted data collection system. 3.

  12. Current distribution evaluation of dye-sensitized solar cell using HTS-SQUID-based magnetic measurement system

    Energy Technology Data Exchange (ETDEWEB)

    Sakai, Kenji, E-mail: Sakai-k@okayama-u.ac.jp; Tanaka, Kohei; Kiwa, Toshihiko; Tsukada, Keiji

    2016-11-15

    Highlights: • Current distribution and direction of dye-sensitized solar cell (DSSC) was measured. • Electrical current flowing in the indium tin oxide (ITO) glass substrate was uniform. • The distribution of electrical current depended on I–V characteristic. • Current direction changed when the performance of DSSC is low. - Abstract: The current flowing inside a dye-sensitized solar cell (DSSC) was measured using a high-temperature superconductor superconducting quantum interference device (HTS-SQUID)-based magnetic measurement system. Further, a new evaluation method of the DSSC, which is difficult to measure using the conventional method, was investigated to improve the characteristics of the DSSC. The tangential components of the magnetic field generated from the DSSC were measured using two HTS-SQUIDs, and the intensity and direction related to the electrical current were obtained by the measured magnetic field. The DSSCs prepared with different dyes and catalytic substances showed different current-intensity mapping. The current direction was different for the DSSC with low performance. In addition, the current flowing in the ITO layer of the ITO glass substrate was also measured and the results confirmed that it had uniform distribution. These results show that the current mapping and the direction of the electrical current depend on the internal factors of the DSSC, and the detection of the magnetic field distribution generated from it is expected to lead to its new evaluation method.

  13. TESTING CPT SYMMETRY WITH CURRENT AND FUTURE CMB MEASUREMENTS

    Energy Technology Data Exchange (ETDEWEB)

    Li, Si-Yu; Zhang, Xinmin [Theory Division, Institute of High Energy Physics, Chinese Academy of Science, P.O. Box 918-4, Beijing 100049 (China); Xia, Jun-Qing; Li, Hong [Key Laboratory of Particle Astrophysics, Institute of High Energy Physics, Chinese Academy of Science, P.O. Box 918-3, Beijing 100049 (China); Li, Mingzhe, E-mail: xiajq@ihep.ac.cn [Interdisciplinary Center for Theoretical Study, University of Science and Technology of China, Hefei, Anhui 230026 (China)

    2015-02-01

    In this paper, we use the current and future cosmic microwave background (CMB) experiments to test the Charge-Parity-Time Reversal (CPT) symmetry. We consider a CPT-violating interaction in the photon sector L{sub cs}∼p{sub μ}A{sub ν} F-tilde {sup μν}, which gives rise to a rotation of the polarization vectors of the propagating CMB photons. By combining the 9 yr WMAP, BOOMERanG 2003, and BICEP1 observations, we obtain the current constraint on the isotropic rotation angle α-bar =−2.12±1.14 (1σ), indicating that the significance of the CPT violation is about 2σ. Here, we particularly take the systematic errors of CMB measurements into account. Then, we study the effects of the anisotropies of the rotation angle [Δα( n-hat )] on the CMB polarization power spectra in detail. Due to the small effects, the current CMB polarization data cannot constrain the related parameters very well. We obtain the 95% C.L. upper limit of the variance of the anisotropies of the rotation angle C {sup α}(0) < 0.035 from all of the CMB data sets. More interestingly, including the anisotropies of rotation angle could lower the best-fit value of r and relax the tension on the constraints of r between BICEP2 and Planck. Finally, we investigate the capabilities of future Planck polarization measurements on α-bar and Δα( n-hat ). Benefited from the high precision of Planck data, the constraints of the rotation angle can be significantly improved.

  14. Clinical use ofa portable electronic device to measure haematocrit

    African Journals Online (AJOL)

    '(BEM), was developed to assist cliriicians to distinguish patients with extreme blood loss from those with normal packed cell volumes. The BEM operates on the princi- ple that blood cells do not conduct a low-frequency alternating current while plasma conducts electricity. Early detection of changes in packed cell volume by.

  15. Pulse Oximeter Derived Blood Pressure Measurement in Patients With a Continuous Flow Left Ventricular Assist Device.

    Science.gov (United States)

    Hellman, Yaron; Malik, Adnan S; Lane, Kathleen A; Shen, Changyu; Wang, I-Wen; Wozniak, Thomas C; Hashmi, Zubair A; Munson, Sarah D; Pickrell, Jeanette; Caccamo, Marco A; Gradus-Pizlo, Irmina; Hadi, Azam

    2017-05-01

    Currently, blood pressure (BP) measurement is obtained noninvasively in patients with continuous flow left ventricular assist device (LVAD) by placing a Doppler probe over the brachial or radial artery with inflation and deflation of a manual BP cuff. We hypothesized that replacing the Doppler probe with a finger-based pulse oximeter can yield BP measurements similar to the Doppler derived mean arterial pressure (MAP). We conducted a prospective study consisting of patients with contemporary continuous flow LVADs. In a small pilot phase I inpatient study, we compared direct arterial line measurements with an automated blood pressure (ABP) cuff, Doppler and pulse oximeter derived MAP. Our main phase II study included LVAD outpatients with a comparison between Doppler, ABP, and pulse oximeter derived MAP. A total of five phase I and 36 phase II patients were recruited during February-June 2014. In phase I, the average MAP measured by pulse oximeter was closer to arterial line MAP rather than Doppler (P = 0.06) or ABP (P < 0.01). In phase II, pulse oximeter MAP (96.6 mm Hg) was significantly closer to Doppler MAP (96.5 mm Hg) when compared to ABP (82.1 mm Hg) (P = 0.0001). Pulse oximeter derived blood pressure measurement may be as reliable as Doppler in patients with continuous flow LVADs. © 2016 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

  16. A simple laser-based device for simultaneous microbial culture and absorbance measurement

    Science.gov (United States)

    Abrevaya, X. C.; Cortón, E.; Areso, O.; Mauas, P. J. D.

    2013-07-01

    In this work we present a device specifically designed to study microbial growth with several applications related to environmental microbiology and other areas of research as astrobiology. The Automated Measuring and Cultivation device (AMC-d) enables semi-continuous absorbance measurements directly during cultivation. It can measure simultaneously up to 16 samples. Growth curves using low and fast growing microorganism were plotted, including Escherichia coli and Haloferax volcanii, a halophilic archaeon.

  17. A Simple Laser-Based Device for Simultaneous Microbial Culture and Absorbance Measurement

    OpenAIRE

    Abrevaya, X. C.; Cortón, E.; Areso, O.; Mauas, P. J. D.

    2012-01-01

    In this work we present a device specifically designed to study microbial growth with several applications related to environmental microbiology and other areas of research as astrobiology. The Automated Measuring and Cultivation device (AMC-d) enables semi-continuous absorbance measurements directly during cultivation. It can measure simultaneously up to 16 samples. Growth curves using low and fast growing microorganism were plotted, including Escherichia coli and Haloferax volcanii, a halop...

  18. 27 CFR 19.277 - Measuring devices and proofing instruments.

    Science.gov (United States)

    2010-04-01

    ... proof or volume. (b) Instruments. Hydrometers and thermometers used by proprietors to gauge spirits... made in conjunction with the volumetric measurement of spirits by meter. If a meter does not have a...

  19. Method and device for measuring formation characteristics of geological formations

    International Nuclear Information System (INIS)

    Antkiw, S.; Murphy, R.D.

    1981-01-01

    A well-logging system is described which uses a pulsed neutron source and which by combining measurements of gamma spectra and neutron characteristics enables such parameters as salinity, porosity, water saturation, lithology and schistosity to be registered directly. (JIW)

  20. A pilot study of a new spectrophotometry device to measure tissue oxygen saturation.

    Science.gov (United States)

    Abel, Gemma; Allen, John; Drinnan, Michael

    2014-09-01

    Tissue oxygen saturation (SO2) measurements have the potential for far wider use than at present but are limited by device availability and portability for many potential applications. A device based on a small, low-cost general-purpose spectrophotometer (the Harrison device) might facilitate wider use. The aim of this study was to compare the Harrison device with a commercial instrument, the LEA O2C.Measurements were carried out on the forearm and finger of 20 healthy volunteers, using a blood pressure cuff on the upper arm to induce different levels of oxygenation. Repeatability of both devices was assessed, and the Bland-Altman method was used to assess agreement between them.The devices showed agreement in overall tracking of changes in SO2. Test-retest agreement for the Harrison device was worse than for O2C, with SD repeatability of 10.6% (forearm) or 18.6% (finger). There was no overall bias between devices, but mean (SD) difference of 1.2 (11.8%) (forearm) or 4.4 (11.5%) (finger) were outside of a clinically acceptable range.Disagreements were attributed to the stability of the Harrison probe and the natural SO2 variations across the skin surface increasing the random error. Therefore, though not equivalent to the LEA O2C, a probe redesign and averaged measurements may help establish the Harrison device as a low cost alternative.

  1. A pilot study of a new spectrophotometry device to measure tissue oxygen saturation

    International Nuclear Information System (INIS)

    Abel, Gemma; Allen, John; Drinnan, Michael

    2014-01-01

    Tissue oxygen saturation (SO2) measurements have the potential for far wider use than at present but are limited by device availability and portability for many potential applications. A device based on a small, low-cost general-purpose spectrophotometer (the Harrison device) might facilitate wider use. The aim of this study was to compare the Harrison device with a commercial instrument, the LEA O2C. Measurements were carried out on the forearm and finger of 20 healthy volunteers, using a blood pressure cuff on the upper arm to induce different levels of oxygenation. Repeatability of both devices was assessed, and the Bland–Altman method was used to assess agreement between them. The devices showed agreement in overall tracking of changes in SO2. Test–retest agreement for the Harrison device was worse than for O2C, with SD repeatability of 10.6% (forearm) or 18.6% (finger). There was no overall bias between devices, but mean (SD) difference of 1.2 (11.8%) (forearm) or 4.4 (11.5%) (finger) were outside of a clinically acceptable range. Disagreements were attributed to the stability of the Harrison probe and the natural SO2 variations across the skin surface increasing the random error. Therefore, though not equivalent to the LEA O2C, a probe redesign and averaged measurements may help establish the Harrison device as a low cost alternative. (paper)

  2. Direct observation of the leakage current in epitaxial diamond Schottky barrier devices by conductive-probe atomic force microscopy and Raman imaging

    OpenAIRE

    Alvarez, Jose; Boutchich, M.; Kleider, J. P.; Teraji, T.; Koide, Y.

    2014-01-01

    The origin of the high leakage current measured in several vertical-type diamond Schottky devices is conjointly investigated by conducting probe atomic force microscopy (CP-AFM) and confocal micro-Raman/Photoluminescence (PL) imaging analysis. Local areas characterized by a strong decrease of the local resistance (5-6 orders of magnitude drop) with respect to their close surrounding have been identified in several different regions of the sample surface. The same local areas, also referenced ...

  3. Reference-Frame-Independent and Measurement-Device-Independent Quantum Key Distribution Using One Single Source

    Science.gov (United States)

    Li, Qian; Zhu, Changhua; Ma, Shuquan; Wei, Kejin; Pei, Changxing

    2018-04-01

    Measurement-device-independent quantum key distribution (MDI-QKD) is immune to all detector side-channel attacks. However, practical implementations of MDI-QKD, which require two-photon interferences from separated independent single-photon sources and a nontrivial reference alignment procedure, are still challenging with current technologies. Here, we propose a scheme that significantly reduces the experimental complexity of two-photon interferences and eliminates reference frame alignment by the combination of plug-and-play and reference frame independent MDI-QKD. Simulation results show that the secure communication distance can be up to 219 km in the finite-data case and the scheme has good potential for practical MDI-QKD systems.

  4. Characterisation of a micro-plasma device sensor using electrical measurements and emission spectroscopy

    International Nuclear Information System (INIS)

    Mariotti, D.

    2002-04-01

    This thesis reports on research undertaken on the characterisation of a micro-plasma device to be used for gas analysis by mean of plasma emission spectroscopy. The work covers aspects related to the micro-plasma electrical and optical emission parameters, and their importance for the utilisation of the micro-plasma device in gas analysis. Experimental results have been used to analyse the fundamental micro-plasma processes and to develop a model, which could provide additional information. This dissertation contains a general literature review of topics related to plasma physics, plasma emission spectroscopy, gas analysis (chemical analysis and artificial olfaction) and other micro-plasma applications. Experimental work focuses on two main areas: electrical measurements and emission measurements. Firstly, electrical measurements are taken and interpretations are given. Where necessary, new theoretical treatments are suggested in order to describe better the physical phenomena. Plasma emission has been considered under different working conditions. This allowed the characterisation of the micro-plasma emission and also a better understanding of the micro-plasma processes. On the basis of the experimental data obtained and other assumptions a model has been developed. A computer simulation based on this model provided additional useful information on the micro- plasma behaviour. The first fundamental implication of this new research is the peculiar behaviour of the micro-plasma. This micro-plasma exhibited deviations from Paschen law and strong dependency on cathode material, which contributed to the formation of a low current stable regime. These results have been followed by physical interpretations and theoretical descriptions. The second implication is the establishment of the boundaries and of the influencing parameters for plasma emission spectroscopy as an analytical tool in this particular micro-plasma. From the applied perspective this study has shown that

  5. Validating a new device for measuring tear evaporation rates.

    Science.gov (United States)

    Rohit, Athira; Ehrmann, Klaus; Naduvilath, Thomas; Willcox, Mark; Stapleton, Fiona

    2014-01-01

    To calibrate and validate a commercially available dermatology instrument to measure tear evaporation rate of contact lens wearers. A dermatology instrument was modified by attaching a swim goggle cup such that the cup sealed around the eye socket. Results for the unmodified instrument are dependent on probe area and enclosed volume. Calibration curves were established using a model eye, to account for individual variations in chamber volume and exposed area. Fifteen participants were recruited and the study included a contact lens wear and a no contact lens wear stage. Day and diurnal variation of the measurements were assessed by taking the measurement three times a day over 2 days. The coefficient of repeatability of the measurement was calculated and a linear mixed model assessed the influence of humidity, temperature, contact lens wear, day and diurnal variations on tear evaporation rate. The associations between variables were assessed using Pearson correlation coefficient. Absolute evaporation rates with and without contact lens wear were calculated based on the new calibration. The measurements were most repeatable during the evening with no lens wear (COR = 49 g m⁻² h) and least repeatable during the evening with contact lens wear (COR = 93 g m⁻² h). Humidity (p = 0.007), and contact lens wear (p evaporation rate. However, temperature (p = 0.54) diurnal variation (p = 0.85) and different days (p = 0.65) had no significant effect after controlling for humidity. Tear evaporation rates can be measured using a modified dermatology instrument. Measurements were higher and more variable with lens wear consistent with previous literature. Control of environmental conditions is important as a higher humidity results in a reduced evaporation rate. © 2013 The Authors Ophthalmic & Physiological Optics © 2013 The College of Optometrists.

  6. Comparison of the methods for determination of calibration and verification intervals of measuring devices

    Directory of Open Access Journals (Sweden)

    Toteva Pavlina

    2017-01-01

    Full Text Available The paper presents different determination and optimisation methods for verification intervals of technical devices for monitoring and measurement based on the requirements of some widely used international standards, e.g. ISO 9001, ISO/IEC 17020, ISO/IEC 17025 etc., maintained by various organizations implementing measuring devices in practice. Comparative analysis of the reviewed methods is conducted in terms of opportunities for assessing the adequacy of interval(s for calibration of measuring devices and their optimisation accepted by an organization – an extension or reduction depending on the obtained results. The advantages and disadvantages of the reviewed methods are discussed, and recommendations for their applicability are provided.

  7. Study of lower hybrid current drive system in tokamak fusion devices

    International Nuclear Information System (INIS)

    Maebara, Sunao

    2001-01-01

    This report describes R and D of a high-power klystron, RF vacuum window, low-outgassing antenna and a front module for a plasma-facing antenna aiming the 5 GHz Lower Hybrid Current Drive (LHCD) system for the next Tokamak Fusion Device. 5 GHz klystron with a low-perveances of 0.7 μP is designed for a high-power and a high-efficiency, the output-power of 715 kW and the efficiency of 63%, which are beyond the conventional design scaling of 450 kW-45%, are performed using the prototype klystron which operates at the pulse duration of 15 μsec. A new pillbox window, which has an oversized length in both the axial and the radial direction, are designed to reduce the RF power density and the electric field strength at the ceramics. It is evaluated that the power capability by cooling edge of ceramics is 1 MW with continuous-wave operation. The antenna module using Dispersion Strengthened Copper which combines high mechanical property up to 500degC with high thermal conductivity, are developed for a low-outgassing antenna in a steady state operation. It is found that the outgassing rate is in the lower range of 4x10 -6 Pam 3 /sm 2 at the module temperature of 300degC, which requires no active vacuum pumping of the LHCD antenna. A front module using Carbon Fiber Composite (CFC) are fabricated and tested for a plasma facing antenna which has a high heat-resistive. Stationary operation of the CFC module with water cooling is performed at the RF power of 46 MWm -2 (about 2 times higher than the design value) during 1000 sec, it is found that the outgassing rate is less than 10 -5 Pam 3 /sm 2 which is low enough for an antenna material. (author)

  8. Measuring device for the coolant flowrate in a reactor core

    International Nuclear Information System (INIS)

    Sawa, Toshihiko.

    1983-01-01

    Purpose: To improve the operation performance by enabling direct and accurate measurement for the reactor core recycling flowrate. Constitution: A control rod guide is disposed to the upper end of a control rod drive mechanism housing passing through the bottom of a reactor pressure vessel and it is inserted into the through hole of a reactor core support plate. A water flow passage is formed through the reactor core support plate for the flowrate measurement of coolants recycled within the reactor core. The static pressure difference between the upper and the lower sides of the reactor core support plate is measured by a pressure difference detector of a pressure difference measuring mechanism, and an output signal from the pressure different detector is inputted to a calculation means, in which the amount of the coolants passing through the water flow passage is calculated based on the output signal corresponding to the pressure difference. Then, the total recycling flowrate in the reactor core is determined in the calculation means based on the relation between the measured flowrate and a predetermined total reactor core recycling flowrate. (Horiuchi, T.)

  9. On-line display used with cathode ray tube film measuring device

    International Nuclear Information System (INIS)

    Fortney, L.R.; Robertson, W.J.

    1981-01-01

    An improved display has been developed for use on our computer controlled measuring device (RIPPLE). The device features a television image of the film and a digital presentation on the same X, Y display. The television image is formed using a modified left and right raster scan which can cover 50% more area in the same time as the traditional raster

  10. Measurement campaign on connectivity of mesh networks formed by mobile devices

    DEFF Research Database (Denmark)

    Pietrarca, Beatrice; Sasso, Giovanni; Perrucci, Gian Paolo

    2007-01-01

    This paper reports the results of a measurement campaign on the connectivity level of mobile devices using Bluetooth (BT) to form cooperative mobile mesh networks. Such mobile mesh networks composed of mobile devices are the basis for any peer-to-peer communication like wireless grids or social...

  11. A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study

    NARCIS (Netherlands)

    Schoot, T.S.; Weenk, M.; Belt, T.H. van de; Engelen, L.J.L.P.G.; Goor, H. van; Bredie, S.J.H.

    2016-01-01

    BACKGROUND: Cuffless blood pressure (BP) monitoring devices, based on pulse transit time, are being developed as an easy-to-use, more convenient, fast, and relatively cheap alternative to conventional BP measuring devices based on cuff occlusion. Thereby they may provide a great alternative to BP

  12. Methods and devices used to measure friction in rolling

    DEFF Research Database (Denmark)

    Jeswiet, J.; Arentoft, Mogens; Henningsen, Poul

    2006-01-01

    Friction at the workpiece-die boundary, in both bulk forming and sheet forming is, arguably, the single most important physical parameter influencing the processing of metals; yet it remains the least understood. Hence there is a need for basic research into metal-die interface mechanisms. To gain...... a good understanding of the mechanisms at the interface and to be able to verify the friction and tribology models that exist, friction sensors are needed. Designing sensors to measure frictional stress in metal working has been pursued by many researchers. This paper surveys methods that have been used...... to measure friction in rolling in the past and discusses some of the recent sensor designs that can now be used to measure friction both in production situations and for research purposes....

  13. RF-assisted current startup in the fusion engineering device (FED)

    International Nuclear Information System (INIS)

    Borowski, S.K.; Peng, Y-K.M.; Kammash, T.

    1982-01-01

    Auxiliary rf heating of electrons before and during the current rise phase in FED is examined as a means of reducing both the initiation loop voltage and resistive flux expenditure during startup. Prior to current initiation 1 to 2 MW of electron cyclotron resonance heating (ECRH) power at --90 GHz is used to create a small volume of high conductivity plasma (Tsub(e) asymptoticaly equals 100 eV, nsub(e) asymptoticaly equals 10 13 cm 3 ) near the upper hybrid resonance (UHR) region. This plasma conditioning permits a small radius (asub(o) asymptoticaly equals 0.2 - 0.4 m) current channel to be established with a relatively low initial loop voltage (<=25 V). During the subsequent plasma expansion and current ramp phase, additional rf power is introduced to reduce volt-second consumption due to plasma resistance. A near classical particle and energy transport model has been developed to estimate the efficiency of electron heating in a currentless, toroidal plasma. The model assumes that MHD instabilities are absent due to the presence of an ambipolar sheath potential and a conducting limiter and vacuum vessel which ''short-circuits'' the vertical charge separation. Preferential electron heating at the UHR leads to the formation of an ambipolar electric field (Esub( a m b)) at the conducting vessel and limiter, which introduces an effective rotational transform via poloidal E vectorsub( a m b) x B vector drift. This drift improves particle confinement and enables the plasma to neutralize itself. The benefits of this effective electrostatic confinement are tempered; however, by the possibility of significant secondary electron emission from the limiters and vessel wall. Reasonable good agreement has been found between our theoretical estimates and the measurements made during ECR preheating experiments on the ISX-B tokamak. This agreement provides some confidence in the preheating power estimates obtained for the FED. (author)

  14. Enhancing the performance of the measurement-device-independent quantum key distribution with heralded pair-coherent sources

    Energy Technology Data Exchange (ETDEWEB)

    Zhu, Feng; Zhang, Chun-Hui; Liu, Ai-Ping [Institute of Signal Processing Transmission, Nanjing University of Posts and Telecommunications, Nanjing 210003 (China); Key Lab of Broadband Wireless Communication and Sensor Network Technology, Nanjing University of Posts and Telecommunications, Ministry of Education, Nanjing 210003 (China); Wang, Qin, E-mail: qinw@njupt.edu.cn [Institute of Signal Processing Transmission, Nanjing University of Posts and Telecommunications, Nanjing 210003 (China); Key Lab of Broadband Wireless Communication and Sensor Network Technology, Nanjing University of Posts and Telecommunications, Ministry of Education, Nanjing 210003 (China); Key Laboratory of Quantum Information, University of Science and Technology of China, Hefei 230026 (China)

    2016-04-01

    In this paper, we propose to implement the heralded pair-coherent source into the measurement-device-independent quantum key distribution. By comparing its performance with other existing schemes, we demonstrate that our new scheme can overcome many shortcomings existing in current schemes, and show excellent behavior in the quantum key distribution. Moreover, even when taking the statistical fluctuation into account, we can still obtain quite high key generation rate at very long transmission distance by using our new scheme. - Highlights: • Implement the heralded pair-coherent source into the measurement-device-independent quantum key distribution. • Overcome many shortcomings existing in current schemes and show excellent behavior. • Obtain quite high key generation rate even when taking statistical fluctuation into account.

  15. Localized corrosion information using high resolution measurement devices

    DEFF Research Database (Denmark)

    Ambat, Rajan

    2005-01-01

    High performance demand for several engineering alloys and components, and miniaturization of electronics and development of MEMS requires better understanding of local corrosion characteristics frequently down to µm scale. This is because in metallic materials corrosion is a sensitive function...... in conjunction with microstructural analysis, using advanced microscopic tools, becomes very important. Corrosion of microelectronics circuits and MEMs is also a recent problem, which demands measurement resolution down to few microns as the components are extremely small, and measurement needs to be carried out...

  16. Measurement of charm in charged current at HERA

    Energy Technology Data Exchange (ETDEWEB)

    Zimmermann, Tobias

    2008-12-15

    A measurement of charm production in charged current (CC) polarized electron-proton deep inelastic scattering processes with data from the H1 detector at the HERA collider is presented. This process in principle allows access to the strange quark density in the proton. In total 5460 CC candidate events in e{sup +}p and 6253 in e{sup -}p data are selected in the kinematic range Q{sup 2}>223 GeV{sup 2} and 0.03measured CC cross sections are {sigma}{sub CC}=(28.9{+-} 1.4)+P{sub e}.(28.6{+-}4.7) pb for e{sup +}p and {sigma}{sub CC}=(49.2{+-}2.3)-P{sub e}.(42.5 {+-}6.8) pb for e{sup -}p, where P{sub e} is the lepton beam polarization. While the measured cross section for e{sup +}p data is in agreement with the theoretical prediction, the cross section for e{sup -}p data shows a weaker dependence on P{sub e} than predicted. The charm fractions in the selected CC candidate event samples are extracted using the muon charge asymmetry. Muons originating from charmed hadron decays in CC events at HERA always have the same charge as the beam lepton. The extracted charm fractions in the selected CC candidate event samples are F{sub c}=9.5{+-}8.9{+-}3.0 % for e{sup +}p and F{sub c}=4.4{+-}6.9{+-}2.6 % for e{sup -}p. Due to the large statistical errors of the measured charm fractions, the strange quark density in the proton has not been extracted. (orig.)

  17. Clinical use of a portable electronic device to measure haematocrit ...

    African Journals Online (AJOL)

    Mean plasma total protein and albumin concentrations were lower compared with normal reference ranges. Six of the 24 patients were acidotic and 4 alkalotic. Leucocyte counts obtained randomly from 13 patients were elevated. Changes in measurements which could influence conductivity did not affect the BEM reading.

  18. Field evaluation of a novel haemoglobin measuring device ...

    African Journals Online (AJOL)

    Objective. To evaluate the use of a robust, cheap method for haemoglobin estimation by non-laboratory-trained personnel in a rural setting. Design. Comparative study. Setting. Tintswalo Hospital. Acomhoek. Participants. 7 nursing sisters, 4 medical students, 2 lay persons. Outcome measures. Haemoglobin estimates ...

  19. Device for measuring neutron-flux distribution density

    International Nuclear Information System (INIS)

    Rozenbljum, N.D.; Mitelman, M.G.; Kononovich, A.A.; Kirsanov, V.S.; Zagadkin, V.A.

    1977-01-01

    An arrangement is described for measuring the distribution of neutron flux density over the height of a nuclear reactor core and which may be used for monitoring energy release or for detecting deviations of neutron flux from an optimal level so that subsequent balance can be achieved. It avoids mutual interference of detectors. Full constructional details are given. (UK)

  20. Device accurately measures and records low gas-flow rates

    Science.gov (United States)

    Branum, L. W.

    1966-01-01

    Free-floating piston in a vertical column accurately measures and records low gas-flow rates. The system may be calibrated, using an adjustable flow-rate gas supply, a low pressure gage, and a sequence recorder. From the calibration rates, a nomograph may be made for easy reduction. Temperature correction may be added for further accuracy.

  1. Device for measuring the dose rate of pulsed neutrons

    International Nuclear Information System (INIS)

    Klett, A.

    2009-01-01

    The author presents a new apparatus, developed in collaboration by Berthold Technologies and the German company DESY, allowing neutron pulsed fields to be measured. It is based on the activation by high energy neutrons of carbon 12 present in the sensor materials, and on the decay of short life radionuclides produced by this activation. The detection principle and system are briefly presented

  2. History and current safety measures at Laguna Palcacocha, Huaraz, Peru

    Science.gov (United States)

    Salazar Checa, César; Cochachin, Alejo; Frey, Holger; Huggel, Christian; Portocarrero, César

    2017-04-01

    Laguna Palcacocha is a large glacier lake in the Cordillera Blanca, Peru, located in the Quillcay catchment, above the city of Huaraz, the local capital. On 13 December 1941, the moraine dam lake collapsed, probably after having been impacted by a large ice avalanche, and triggered a major outburst flood. This GLOF destroyed about a third of the city of Huaraz, causing about 2,000 casualties and is therefore one of the deadliest glacier lake outbursts known in history. In 1974, the Glaciology Unit of Peru, responsible for the studying, monitoring and mitigation works related to glacier hazards installed a reinforcement of the natural moraine dam of the newly filled Laguna Palcacocha, with an artificial drainage channel at 7 m below the crest of the reinforced dam. At that time, the lake had an area of 66,800 m2 and a volume of 0.5 x 106 m3. During the past decades, in the course of continued glacier retreat, Laguna Palcacocha has undergone an extreme growth. In February 2016, the lake had an area of 514,000 m2 (7.7 times the area of 1974) and a volume of more than 17 x 106 m3 (more than 34 times the volume of 1974). At the same time, the city of Huaraz, located 20 km downstream of the lake, grew significantly after its almost complete destruction by the 1970 earthquake. Today, about 120,000 people are living in the city. Due to the persisting possibility for large ice avalanches directly above the Palcacocha lake, this constitutes a high-risk situation, requiring new hazard and risk mitigation measures. As an immediate temporal measure, in order to bridge the time until the realization of a more permanent measure, a syphoning system has been installed in 2011, using about ten 700-m pipes with a 10-inch (25.4 cm) diameter. The aim of this syphoning attempt is to lower the lake level by about 7 m, and therefore reduce the lake volume on the one hand, and also reach a higher dam freeboard. However, the system is less effective than assumed, currently the lake level

  3. Principle and Control Design of Active Ground-Fault Arc Suppression Device for Full Compensation of Ground Current

    DEFF Research Database (Denmark)

    Wang, Wen; Zeng, Xiangjun; Yan, Lingjie

    2017-01-01

    current into the neutral without any large-capacity reactors, and thus avoids the aforementioned overvoltage. It compensates all the active, reactive and harmonic components of the ground current to reliably extinguish the ground-fault arcs. A dual-loop voltage control method is proposed to realize arc...... suppression without capacitive current detection. Its time-based feature also brings the benefit of fast response on ground-fault arc suppression. The principle of full current compensation is analyzed, together with the controller design method of the proposed device. Experiment on a prototype was carried...

  4. Design and testing of an innovative solar radiation measurement device

    International Nuclear Information System (INIS)

    Badran, Omar; Al-Salaymeh, Ahmed; El-Tous, Yousif; Abdala, Wasfi

    2010-01-01

    After review of studies conducted on the solar radiation measuring systems, a new innovative instrument that would help in measuring the accurate solar radiation on horizontal surfaces has been designed and tested. An advanced instrument with ease of use and high precision that would enable the user to take the readings in terms of solar intensity (W/m 2 ) has been tested. Also, the innovative instrument can record instantaneous readings of the solar intensities as well as the averages value of the solar radiation flux during certain periods of time. The instrument based in its design on being programmed by programmable interfacing controller (PIC). Furthermore, the power supply circuit is fed by the solar energy cells and does not need an external power source.

  5. Mean-Square Error Due to Gradiometer Field Measuring Devices

    Science.gov (United States)

    1991-06-01

    convolving the gradiometer data with the inverse transform of I /T(a, 13), applying an ap- Hence (2) may be expressed in the transform domain as propriate... inverse transform of I / T(ot, 1) will not be possible quency measurements," Superconductor Applications: SQUID’s and because its inverse does not exist...and because it is a high- Machines, B. B. Schwartz and S. Foner, Eds. New York: Plenum pass function its use in an inverse transform technique Press

  6. Awareness training and hearing protection devices: Current practices in the South African mining industry

    CSIR Research Space (South Africa)

    Edwards, A

    2012-09-01

    Full Text Available This presentation outlines the importance of awareness training of managers at all levels and miners regarding the importance of hearing protection devices and adequate knowledge, motivation and training to prevent hearing loss....

  7. An electrochemical sensor device for measuring blood ammonia at the point of care.

    Science.gov (United States)

    Brannelly, N T; Killard, A J

    2017-05-15

    The level of ammonia in blood is relevant in a number of medical conditions. While ammonia is a marker of dysfunction, elevated ammonia is itself a serious medical emergency and can lead to significant and permanent neurological impairment if not addressed quickly. Blood ammonia testing is typically performed in the central laboratory. While a number of point of care devices have been developed, these are based on classical enzymatic or colorimetric principles and have not been widely adopted. In this work, an electrochemical sensor device was developed for measuring blood ammonia. The device was based on the deposition of polyaniline nanoparticle films onto screen printed interdigitated electrodes using inkjet printing and their integration into a polymer microfabricated device with a polytetrafluoroethylene membrane. The device required a 52µL serum sample and measured the change in impedance of the sensor with respect to air at 1kHz, 5mV rms. The device was capable of the measurement of ammonia in serum across the physiologically relevant range of 25-200µM (r 2 =0.9984) and had a limit of detection of 12µM (n =3). The device showed no significant issues with common electrochemical interferences in blood. The device was also validated against a commercial spectrophotometric assay which resulted in excellent correlation (r =0.9699, pair (n =12) and could be stored in desiccant for at least five months. Copyright © 2017 Elsevier B.V. All rights reserved.

  8. Measurement of charm in charged current at HERA

    International Nuclear Information System (INIS)

    Zimmermann, Tobias

    2008-12-01

    A measurement of charm production in charged current (CC) polarized electron-proton deep inelastic scattering processes with data from the H1 detector at the HERA collider is presented. This process in principle allows access to the strange quark density in the proton. In total 5460 CC candidate events in e + p and 6253 in e - p data are selected in the kinematic range Q 2 >223 GeV 2 and 0.03 CC =(28.9± 1.4)+P e .(28.6±4.7) pb for e + p and σ CC =(49.2±2.3)-P e .(42.5 ±6.8) pb for e - p, where P e is the lepton beam polarization. While the measured cross section for e + p data is in agreement with the theoretical prediction, the cross section for e - p data shows a weaker dependence on P e than predicted. The charm fractions in the selected CC candidate event samples are extracted using the muon charge asymmetry. Muons originating from charmed hadron decays in CC events at HERA always have the same charge as the beam lepton. The extracted charm fractions in the selected CC candidate event samples are F c =9.5±8.9±3.0 % for e + p and F c =4.4±6.9±2.6 % for e - p. Due to the large statistical errors of the measured charm fractions, the strange quark density in the proton has not been extracted. (orig.)

  9. Poster Abstract: Automatic Calibration of Device Attitude in Inertial Measurement Unit Based Traffic Probe Vehicles

    KAUST Repository

    Mousa, Mustafa; Sharma, Kapil; Claudel, Christian

    2016-01-01

    to replace them with inertial measurement units onboard vehicles, to estimate vehicle location and attitude using inertial data only. While promising, this technology requires one to carefully calibrate the orientation of the device inside the vehicle

  10. Development path and current status of the NANIVID: a new device for cancer cell studies

    Science.gov (United States)

    Raja, Waseem Khan; Padgen, Michael R.; Williams, James K.; Wyckoff, Jeffrey; Condeelis, John; Castracane, James

    2011-02-01

    Cancer cells create a unique microenvironment in vivo which enables migration to distant organs. To better understand the tumor microenvironment, special tools and devices are required to monitor the interactions between different cell types and the effects of particular chemical gradients. This study presents the design and optimization of a new, versatile chemotaxis device called the NANIVID (NANo IntraVital Device). The device is fabricated using BioMEMS techniques and consists of etched and bonded Pyrex substrates, a soluble factor reservoir, fluorescent tracking beads and a microelectrode array for cell quantification. The reservoir contains a customized hydrogel blend loaded with EGF which diffuses out of the hydrogel to create a chemotactic gradient. This reservoir sustains a steady release of growth factor into the surrounding environment for many hours and establishes a concentration gradient that attracts specific cells to the device. In addition to a cell collection tool, the NANIVID can be modified to act as a delivery vehicle for the local generation of alternate soluble factor gradients to initiate controlled changes to the microenvironment such as hypoxia, ECM stiffness and etc. The focus of this study is to design and optimize the new device for wide ranging studies of breast cancer cell dynamics in vitro and ultimately, implantation for in vivo work.

  11. Development of an Automated Gear Tooth Contour Measuring Device.

    Science.gov (United States)

    1982-04-12

    even under the Influence of body beatL " and changng air temperature. Operating the machine in the fluid-filled mode greatly 5-1... the finaL Phase I report incorporating draft corrections. 19. Key WRDS (Coafw so.. me Mo od*f se**p ow somse W, 610 mA..) Mensuration Coordinate...Measuring Machine Noncontact Optic Probe Surface Wear Assessment 2L &911 RACY (Coneaa Me me... odds N msoosw mW SOMWI IV Wle"m " This report covers the

  12. The wave vane - A device to measure the breaker angle

    Digital Repository Service at National Institute of Oceanography (India)

    Chandramohan, P.; Nayak, B.U.; Anand, N.M.

    directional wave gauge at a considerable cost. For economic and practical considerations, visual measurements on the breaker height, the breaker period and the breaker angle are generally made in the LEO pro- gramme. The breaker height is recorded... of the fins. Flag masts are fixed on either end of the arms. The whole assembly thus revolves around the spindle resting on the sleeve. WORKING PRINCIPLE The flag masts, two on the revolving arms and one on the top of the spindle lie always in a straight...

  13. Shaf rotation speed measurement device, its checking and variations checking

    International Nuclear Information System (INIS)

    Gadrault, Robert.

    1976-01-01

    Appliance for measuring the rotational speed of a shaft and monitoring of this speed and its changes. The uses to be made specifically concern the nuclear field and in this field the drive shafts of water coolant feed pumps. Detecting te rotation of the shaft concerned may be resolved with electronic sensors or proximity detectors which, because they are not in mechanical connexion with the shaft the speed of which they are to help detect, do not bring any lack of precision. The accuracy of the determination them depends only on the downstream processing of the data supplied by the sensor [fr

  14. Time measurements with a mobile device using sound

    Science.gov (United States)

    Wisman, Raymond F.; Spahn, Gabriel; Forinash, Kyle

    2018-05-01

    Data collection is a fundamental skill in science education, one that students generally practice in a controlled setting using equipment only available in the classroom laboratory. However, using smartphones with their built-in sensors and often free apps, many fundamental experiments can be performed outside the laboratory. Taking advantage of these tools often require creative approaches to data collection and exploring alternative strategies for experimental procedures. As examples, we present several experiments using smartphones and apps that record and analyze sound to measure a variety of physical properties.

  15. Device for measuring flow rate in a nuclear reactor core

    International Nuclear Information System (INIS)

    Hamano, Jiro.

    1980-01-01

    Purpose: To always calculate core flow rate automatically and accurately in BWR type nuclear power plants. Constitution: Jet pumps are provided to the recycling pump and to the inside of the pressure vessel of a nuclear reactor. The jet pumps comprise a plurality of calibrated jet pumps for forcively convecting the coolants and a plurality of not calibrated jet pumps in order to cool the heat generated in the reactor core. The difference in the pressures between the upper and the lower portions in both of the jet pumps is measured by difference pressure transducers. Further, a thermo-sensitive element is provided to measure the temperature of recycling water at the inlet of the recycling pump. The output signal from the difference pressure transducer is inputted to a process computer, calculated periodically based on predetermined calculation equations, compensated for the temperature by a recycling water temperature signal and outputted as a core flow rate signal to a recoder. The signal is also used for the power distribution calculation in the process computer and the minimum limit power ratio as the thermal limit value for the fuels is outputted. (Furukawa, Y.)

  16. Device for measuring high temperature heat conductivity of solids and melts

    International Nuclear Information System (INIS)

    Magomedov, Ya.B.; Gadzhiev, G.G.

    1990-01-01

    A modification of a device for measuring heat conductivity by a compensation method when a thermocouple with gadolinium sulfide being used is suggested. Such a device has less error of measurement (8%), wider interval of working temperatures (300-1600K) and it permits to investigate the material in the wide range of heat conductivity values (0.5-30 W/(mxK)). The stainless steel 12Kh18N10T, lanthanum sulfide and melted quartz were used for the device calibration. The results obtained and the literature data on these materials agree well between each other

  17. Light collection optics for measuring flux and spectrum from light-emitting devices

    Science.gov (United States)

    McCord, Mark A.; DiRegolo, Joseph A.; Gluszczak, Michael R.

    2016-05-24

    Systems and methods for accurately measuring the luminous flux and color (spectra) from light-emitting devices are disclosed. An integrating sphere may be utilized to directly receive a first portion of light emitted by a light-emitting device through an opening defined on the integrating sphere. A light collector may be utilized to collect a second portion of light emitted by the light-emitting device and direct the second portion of light into the integrating sphere through the opening defined on the integrating sphere. A spectrometer may be utilized to measure at least one property of the first portion and the second portion of light received by the integrating sphere.

  18. Study of lower hybrid current drive system in tokamak fusion devices

    Energy Technology Data Exchange (ETDEWEB)

    Maebara, Sunao [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    2001-01-01

    This report describes R and D of a high-power klystron, RF vacuum window, low-outgassing antenna and a front module for a plasma-facing antenna aiming the 5 GHz Lower Hybrid Current Drive (LHCD) system for the next Tokamak Fusion Device. 5 GHz klystron with a low-perveances of 0.7 {mu}P is designed for a high-power and a high-efficiency, the output-power of 715 kW and the efficiency of 63%, which are beyond the conventional design scaling of 450 kW-45%, are performed using the prototype klystron which operates at the pulse duration of 15 {mu}sec. A new pillbox window, which has an oversized length in both the axial and the radial direction, are designed to reduce the RF power density and the electric field strength at the ceramics. It is evaluated that the power capability by cooling edge of ceramics is 1 MW with continuous-wave operation. The antenna module using Dispersion Strengthened Copper which combines high mechanical property up to 500degC with high thermal conductivity, are developed for a low-outgassing antenna in a steady state operation. It is found that the outgassing rate is in the lower range of 4x10{sup -6} Pam{sup 3}/sm{sup 2} at the module temperature of 300degC, which requires no active vacuum pumping of the LHCD antenna. A front module using Carbon Fiber Composite (CFC) are fabricated and tested for a plasma facing antenna which has a high heat-resistive. Stationary operation of the CFC module with water cooling is performed at the RF power of 46 MWm{sup -2} (about 2 times higher than the design value) during 1000 sec, it is found that the outgassing rate is less than 10{sup -5} Pam{sup 3}/sm{sup 2} which is low enough for an antenna material. (author)

  19. Development of advanced materials and devices for nuclear radiation measurements

    International Nuclear Information System (INIS)

    Gadkari, S.C.

    2015-01-01

    Single crystals of technologically important materials are grown in the Crystal Technology Section of the Technical Physics Division, BARC. These crystals find applications as scintillators and dosimeters in nuclear radiation detection/measurements. Scintillator crystals of some advanced materials like cerium doped Gd 3 Ga 3 Al 2 O 12 , Lu 2 SiO 5 , YAIO 3 etc and some conventional materials such as Bi 4 Ge 3 O 12 , CsI:Tl, NaI:Tl, etc have been grown from melts using the Czochralski and Bridgman techniques. Portable gamma-ray spectrometers that work from a USB port of a laptop have been developed using the grown scintillator crystals. In recent years there has been a flurry of research activities on materials containing Li 6 , B 10 , etc that have large capture cross-sections for neutrons to develop solid state detectors for neutrons. For this purpose single crystals of cerium doped Li 6 Y(BO 3 ) 3 and silver doped Li 2 B 4 O 7 have been developed. Optical, thermo-luminescence, photo-luminescence and scintillation properties of these crystals have been investigated with a view to develop detectors and dosimeters. The Li 2 B 4 O 7 :Ag is a tissue equivalent material (Z eff = 7.3 close to 7.4 of tissue) useful in the personal and medical dosimetry applications. As the emission of Ag + lies in the UV region (267 nm), a customized TL measurement set-up has been developed using a solar blind PMT that enabled the measurement of very low doses below 5 μGy and linearity up to 100 Gy. Films of CsI:TI in the 10 nm to 3 μm thickness range were deposited on silicon substrates using the physical vapor deposition technique under vacuum conditions. The deposited films investigated using SEM and AFM revealed a columnar growth behavior with a preferential orientation along <200>. The growth of single crystals from melts, recent efforts in the development of detectors and results of experiments conducted to detect thermal neutrons are described. (author)

  20. Measurements of the PLT and PDX device activation

    International Nuclear Information System (INIS)

    Stavely, J.; Barnes, C.W.; Chrien, R.E.; Strachan, J.D.

    1981-09-01

    Measurements of the activation levels around the PLT and PDX tokamaks have been made using a Ge(Li) gamma spectrometer and a Geiger counter. The activation results from radiation induced in the plasma by 14 MeV neutrons from the d(t,n)α fusion reaction, 14.7 MeV protons from the d( 3 He,p)α fusion reaction, 10 → 20 MeV hard x-rays from runaway electron induced bremmstrahlung, and 2.5 MeV neutrons from the d(d,n) 3 He fusion reaction. The magnitude of the activation is compared to that predicted for PDX on the basis of one-dimensional activation codes

  1. Experimental arrangement to measure dispersion in optical fiber devices

    Energy Technology Data Exchange (ETDEWEB)

    Armas Rivera, Ivan [Benemerita Universidad Autonoma de Puebla, Facultad de Ciencias de la Electronica (Mexico); Beltran Perez, Georgina; Castillo Mixcoatl, Juan; Munoz Aguirre, Severino [Benemerita Universidad Autonoma de Puebla, Facultad de Ciencias Fisico Matematicas (Mexico); Zaca Moran, Placido, E-mail: ivan_rr1@hotmail.com [Benemerita Universidad Autonoma de Puebla, Fisicoquimica de Materiales ICUAP (Mexico)

    2011-01-01

    Dispersion is a quite important parameter in systems based on optical fiber, especially in pulsed emission lasers, where the temporal width is affected by such parameter. Therefore, it is necessary to consider the dispersion provoked by each component in the cavity. There are various experimental interferometric arrangements to evaluate this parameter. Generally, these systems modify the wavelength to obtain information about the n({lambda}) dependency, which is contained in the interferogram phase. However, this makes the system quite slow and it requires tunable and narrow bandwidth laser sources. In the present work, results obtained from an arrangement based on Mach-Zehnder interferometer where one of the arms is the optical fiber under study, while the reference one is air, are presented. In order to determine the n({lambda}) dependency, a wide spectrum light source was used in the wavelength range of interest. The phase information was evaluated from the interferometric signal measured by an optical spectrum analyzer.

  2. Experimental arrangement to measure dispersion in optical fiber devices

    International Nuclear Information System (INIS)

    Armas Rivera, Ivan; Beltran Perez, Georgina; Castillo Mixcoatl, Juan; Munoz Aguirre, Severino; Zaca Moran, Placido

    2011-01-01

    Dispersion is a quite important parameter in systems based on optical fiber, especially in pulsed emission lasers, where the temporal width is affected by such parameter. Therefore, it is necessary to consider the dispersion provoked by each component in the cavity. There are various experimental interferometric arrangements to evaluate this parameter. Generally, these systems modify the wavelength to obtain information about the n(λ) dependency, which is contained in the interferogram phase. However, this makes the system quite slow and it requires tunable and narrow bandwidth laser sources. In the present work, results obtained from an arrangement based on Mach-Zehnder interferometer where one of the arms is the optical fiber under study, while the reference one is air, are presented. In order to determine the n(λ) dependency, a wide spectrum light source was used in the wavelength range of interest. The phase information was evaluated from the interferometric signal measured by an optical spectrum analyzer.

  3. Measurement-Device Independency Analysis of Continuous-Variable Quantum Digital Signature

    Directory of Open Access Journals (Sweden)

    Tao Shang

    2018-04-01

    Full Text Available With the practical implementation of continuous-variable quantum cryptographic protocols, security problems resulting from measurement-device loopholes are being given increasing attention. At present, research on measurement-device independency analysis is limited in quantum key distribution protocols, while there exist different security problems for different protocols. Considering the importance of quantum digital signature in quantum cryptography, in this paper, we attempt to analyze the measurement-device independency of continuous-variable quantum digital signature, especially continuous-variable quantum homomorphic signature. Firstly, we calculate the upper bound of the error rate of a protocol. If it is negligible on condition that all measurement devices are untrusted, the protocol is deemed to be measurement-device-independent. Then, we simplify the calculation by using the characteristics of continuous variables and prove the measurement-device independency of the protocol according to the calculation result. In addition, the proposed analysis method can be extended to other quantum cryptographic protocols besides continuous-variable quantum homomorphic signature.

  4. Analysis of strain distribution and critical current of superconductors based on a strain-critical current measurement system

    International Nuclear Information System (INIS)

    Liu Fang; Wu Yu; Long Feng

    2010-01-01

    Based on Pacman device which is widely used to investigate the axial strain dependence of the critical current in superconductors, the finite element analysis method is employed to carry out the force analysis of the spring and the superconducting strand, thereby the axial and lateral strain distributions of the superconducting strand are obtained. According to the two extreme assumptions(low inter-filament resistance and high inter-filament resistance), the effects of the strain homogeneity at the cross section of the superconductor on the critical current is analyzed combined with the Nb 3 Sn deviatoric strain-critical current scaling law. (authors)

  5. Measuring device for strong gamma-ray sources; Dispositif de mesure des fortes sources emettrices {gamma}

    Energy Technology Data Exchange (ETDEWEB)

    Engelman, J; Vagner, J [Commissariat a l' Energie Atomique, Saclay (France). Centre d' Etudes Nucleaires

    1956-07-01

    We are given the description of a hollow argon-filled cylindrical ionisation chamber which is to be used to measure gamma-emitting sources. The instrument is currently used at the Measures Department in routine gauging of some radioelements. Sources are introduced into the central part of the chamber through a remote handling device. Measures are directly registered, it is not worth while removing the source from the container; a deviation of the source has little effect on the ionization current. The chamber was gauged to test such elements as: {sup 198}Au, {sup 60}Co, {sup 192}Ir, {sup 24}Na, {sup 137}Cs. Its measuring power approximately ranges from 100 micro-curies to 5 curies. (author) [French] On decrit une chambre d'ionisation cylindrique creuse, a remplissage d'argon, destinee a la mesure des sources emettrices {gamma}. Cet appareil est utilise couramment par la Section Mesures pour l'etalonnage de routine d'un certain nombre de radioelements. Les sources sont mises en place au centre de la chambre par un dispositif de manipulation a distance. La mesure est faite directement, sans qu'il soit necessaire d'extraire la source de son container; un decentrement de la source n'a en effet pas d'influence sensible sur le courant d'ionisation. Cette chambre d'ionisation a ete etalonnee pour divers radioelements: {sup 198}Au, {sup 60}Co, {sup 192}Ir, {sup 24}Na, {sup 137}Cs. La flamme d'activite mesurable s'etend de 100 microcuries a 5 curies, environ. (auteur)

  6. Waveform measurement in mocrowave device characterization: impact on power amplifiers design

    Directory of Open Access Journals (Sweden)

    Roberto Quaglia

    2016-07-01

    Full Text Available This paper describes an example of a measurement setup enabling waveform measurements during the load-pull characterization of a microwave power device. The significance of this measurement feature is highlighted showing how waveform engineering can be exploited to design high efficiency microwave power amplifiers.

  7. Electron and current density measurements on tokamak plasmas

    International Nuclear Information System (INIS)

    Lammeren, A.C.A.P. van.

    1991-01-01

    The first part of this thesis describes the Thomson-scattering diagnostic as it was present at the TORTUR tokamak. For the first time with this diagnostic a complete tangential scattering spectrum was recorded during one single laser pulse. From this scattering spectrum the local current density was derived. Small deviations from the expected gaussian scattering spectrum were observed indicating the non-Maxwellian character of the electron-velocity distribution. The second part of this thesis describes the multi-channel interferometer/ polarimeter diagnostic which was constructed, build and operated on the Rijnhuizen Tokamak Project (RTP) tokamak. The diagnostic was operated routinely, yielding the development of the density profiles for every discharge. When ECRH (Electron Cyclotron Resonance Heating) is switched on the density profile broadens, the central density decreases and the total density increases, the opposite takes place when ECRH is switched off. The influence of MHD (magnetohydrodynamics) activity on the density was clearly observable. In the central region of the plasma it was measured that in hydrogen discharges the so-called sawtooth collapse is preceded by an m=1 instability which grows rapidly. An increase in radius of this m=1 mode of 1.5 cm just before the crash is observed. In hydrogen discharges the sawtooth induced density pulse shows an asymmetry for the high- and low-field side propagation. This asymmetry disappeared for helium discharges. From the location of the maximum density variations during an m=2 mode the position of the q=2 surface is derived. The density profiles are measured during the energy quench phase of a plasma disruption. A fast flattening and broadening of the density profile is observed. (author). 95 refs.; 66 figs.; 7 tabs

  8. Current measurements over the western continental shelf of India

    Digital Repository Service at National Institute of Oceanography (India)

    Shenoi, S.S.C.; Antony, M.K.

    south during May and March and towards north during November. The spectral analysis indicates that the inertial and semidiurnal frequency motions are the major contributors to the current structure. The inertial currents were concentrated near...

  9. [Regulatory Program for Medical Devices in Cuba: experiences and current challenges].

    Science.gov (United States)

    Pereira, Dulce María Martínez; Rodríguez, Yadira Álvarez; Valdés, Yamila Cedeño; Ribas, Silvia Delgado

    2016-05-01

    Regulatory control of medical devices in Cuba is conducted through a system based on the Regulatory Program for Medical Devices as a way to ensure the safety, efficacy, and effectiveness of these technologies, which are in use by the National Health System. This program was launched in 1992, when the Regulations for State Evaluation and Registration of Medical Devices were approved. Its successive stages and the merging of regulatory activities for drugs and medical equipment have meant progress toward stronger, more transparent strategies and greater control of industry and the National Health System. Throughout its course the Cuban program has met with challenges and difficulties that it has addressed by drawing on its own experiences. During the new period, the greatest challenges revolve around ensuring that regulatory systems incorporate scientific evaluation, risk levels, maximum rigor through the use of technical standards, and the implementation of international recommendations, together with the application of the ISO 13485 certification scheme, enhanced market monitoring, and classification of medical devices in accordance with their relevance to the country's national health policies. From the regional standpoint, the greatest challenge lies in working toward regulatory convergence. The Collaborating Centre for the Regulation of Health Technologies will support the proposed regulatory strategy and established regional priorities, in particular in connection with the implementation of actions involving medical devices.

  10. Measurement of Skin Dose from Using the Treatment Immobilization Devices

    International Nuclear Information System (INIS)

    Je, Jae Yong; Park, Chul Woo; Noh, Kyung Suk

    2009-01-01

    The research was about the relation between the dorsal side dose measured by using the phantom body (Alderson Rando Phantom) and factors like contacted material of the patients, the size of the field, angle of incidence. Compared with mylar (tennis racket), the dose on 10 x 10 cm 2 field size of cotton was increased by 2% and by 8% in the case of breast board. In the case of 15 x 15 cm 2 field size, the dose was increased by 6% compared with 10 x 10 cm 2 size. The field size of 20 x 20 cm 2 resulted in 10% increase of dose, while 5 x 5 cm 2 produced 13% decrease. Compared with incident angle 0 degree, the cases for the incident angle 5 degrees had 0.4% less dose for breast board, 0.5% for tennis racket, 1.1% for cotton. The cases for the incident angle 10 degrees had 1.5% less dose for breast board, 1.9% for tennis racket, 2.6% for cotton. For the incident angle 15 degrees, breast board, tennis racket, cotton caused decrease of dose by 3.9%, 2.6%, 3.86% respectively. Resultantly carbon material can cause more skin dose in treatment field. By the results of this study, we recommend that one should avoid the contact between the carbon material and skin.

  11. Electromagnetic modelling of current flow in the heart from TASER devices and the risk of cardiac dysrhythmias

    Energy Technology Data Exchange (ETDEWEB)

    Holden, S J [Defence Science and Technology Laboratory, Porton Down, Salisbury, Wiltshire SP4 0JQ (United Kingdom); Sheridan, R D [Defence Science and Technology Laboratory, Porton Down, Salisbury, Wiltshire SP4 0JQ (United Kingdom); Coffey, T J [Mathshop Ltd, Porton Down Science Park, Salisbury, Wiltshire SP4 0JQ (United Kingdom); Scaramuzza, R A [Flomerics Ltd, Electromagnetic Division, TLM House, Percy Street, Nottingham NG16 3EP (United Kingdom); Diamantopoulos, P [Bio-Medical Modelling Unit, School of Engineering, University of Sussex, Falmer, Sussex BN1 9QT (United Kingdom)

    2007-12-21

    Increasing use by law enforcement agencies of the M26 and X26 TASER electrical incapacitation devices has raised concerns about the arrhythmogenic potential of these weapons. Using a numerical phantom constructed from medical images of the human body in which the material properties of the tissues are represented, computational electromagnetic modelling has been used to predict the currents arising at the heart following injection of M26 and X26 waveforms at the anterior surface of the chest (with one TASER 'barb' directly overlying the ventricles). The modelling indicated that the peak absolute current densities at the ventricles were 0.66 and 0.11 mA mm{sup -2} for the M26 and X26 waveforms, respectively. When applied during the vulnerable period to the ventricular epicardial surface of guinea-pig isolated hearts, the M26 and X26 waveforms induced ectopic beats, but only at current densities greater than 60-fold those predicted by the modelling. When applied to the ventricles in trains designed to mimic the discharge patterns of the TASER devices, neither waveform induced ventricular fibrillation at peak currents >70-fold (for the M26 waveform) and >240-fold (for the X26) higher than the modelled current densities. This study provides evidence for a lack of arrhythmogenic action of the M26 and X26 TASER devices.

  12. Design of all-optical, hot-electron current-direction-switching device based on geometrical asymmetry.

    Science.gov (United States)

    Kumarasinghe, Chathurangi S; Premaratne, Malin; Gunapala, Sarath D; Agrawal, Govind P

    2016-02-18

    We propose a nano-scale current-direction-switching device(CDSD) that operates based on the novel phenomenon of geometrical asymmetry between two hot-electron generating plasmonic nanostructures. The proposed device is easy to fabricate and economical to develop compared to most other existing designs. It also has the ability to function without external wiring in nano or molecular circuitry since it is powered and controlled optically. We consider a such CDSD made of two dissimilar nanorods separated by a thin but finite potential barrier and theoretically derive the frequency-dependent electron/current flow rate. Our analysis takes in to account the quantum dynamics of electrons inside the nanorods under a periodic optical perturbation that are confined by nanorod boundaries, modelled as finite cylindrical potential wells. The influence of design parameters, such as geometric difference between the two nanorods, their volumes and the barrier width on quality parameters such as frequency-sensitivity of the current flow direction, magnitude of the current flow, positive to negative current ratio, and the energy conversion efficiency is discussed by considering a device made of Ag/TiO2/Ag. Theoretical insight and design guidelines presented here are useful for customizing our proposed CDSD for applications such as self-powered logic gates, power supplies, and sensors.

  13. Optoelectronic device for the measurement of the absolute linear position in the micrometric displacement range

    Science.gov (United States)

    Morlanes, Tomas; de la Pena, Jose L.; Sanchez-Brea, Luis M.; Alonso, Jose; Crespo, Daniel; Saez-Landete, Jose B.; Bernabeu, Eusebio

    2005-07-01

    In this work, an optoelectronic device that provides the absolute position of a measurement element with respect to a pattern scale upon switch-on is presented. That means that there is not a need to perform any kind of transversal displacement after the startup of the system. The optoelectronic device is based on the process of light propagation passing through a slit. A light source with a definite size guarantees the relation of distances between the different elements that constitute our system and allows getting a particular optical intensity profile that can be measured by an electronic post-processing device providing the absolute location of the system with a resolution of 1 micron. The accuracy of this measuring device is restricted to the same limitations of any incremental position optical encoder.

  14. Direct current force sensing device based on compressive spring, permanent magnet, and coil-wound magnetostrictive/piezoelectric laminate.

    Science.gov (United States)

    Leung, Chung Ming; Or, Siu Wing; Ho, S L

    2013-12-01

    A force sensing device capable of sensing dc (or static) compressive forces is developed based on a NAS106N stainless steel compressive spring, a sintered NdFeB permanent magnet, and a coil-wound Tb(0.3)Dy(0.7)Fe(1.92)/Pb(Zr, Ti)O3 magnetostrictive∕piezoelectric laminate. The dc compressive force sensing in the device is evaluated theoretically and experimentally and is found to originate from a unique force-induced, position-dependent, current-driven dc magnetoelectric effect. The sensitivity of the device can be increased by increasing the spring constant of the compressive spring, the size of the permanent magnet, and/or the driving current for the coil-wound laminate. Devices of low-force (20 N) and high-force (200 N) types, showing high output voltages of 262 and 128 mV peak, respectively, are demonstrated at a low driving current of 100 mA peak by using different combinations of compressive spring and permanent magnet.

  15. Sensor-Based Assistive Devices for Visually-Impaired People: Current Status, Challenges, and Future Directions

    Directory of Open Access Journals (Sweden)

    Wafa Elmannai

    2017-03-01

    Full Text Available The World Health Organization (WHO reported that there are 285 million visuallyimpaired people worldwide. Among these individuals, there are 39 million who are totally blind. There have been several systems designed to support visually-impaired people and to improve the quality of their lives. Unfortunately, most of these systems are limited in their capabilities. In this paper, we present a comparative survey of the wearable and portable assistive devices for visuallyimpaired people in order to show the progress in assistive technology for this group of people. Thus, the contribution of this literature survey is to discuss in detail the most significant devices that are presented in the literature to assist this population and highlight the improvements, advantages, disadvantages, and accuracy. Our aim is to address and present most of the issues of these systems to pave the way for other researchers to design devices that ensure safety and independent mobility to visually-impaired people.

  16. Sensor-Based Assistive Devices for Visually-Impaired People: Current Status, Challenges, and Future Directions

    Science.gov (United States)

    Elmannai, Wafa; Elleithy, Khaled

    2017-01-01

    The World Health Organization (WHO) reported that there are 285 million visually-impaired people worldwide. Among these individuals, there are 39 million who are totally blind. There have been several systems designed to support visually-impaired people and to improve the quality of their lives. Unfortunately, most of these systems are limited in their capabilities. In this paper, we present a comparative survey of the wearable and portable assistive devices for visually-impaired people in order to show the progress in assistive technology for this group of people. Thus, the contribution of this literature survey is to discuss in detail the most significant devices that are presented in the literature to assist this population and highlight the improvements, advantages, disadvantages, and accuracy. Our aim is to address and present most of the issues of these systems to pave the way for other researchers to design devices that ensure safety and independent mobility to visually-impaired people. PMID:28287451

  17. High and tunable spin current induced by magnetic-electric fields in a single-mode spintronic device

    International Nuclear Information System (INIS)

    Bala Kumar, S; Jalil, M B A; Tan, S G; Liang, G-C

    2009-01-01

    We proposed that a viable form of spin current transistor is one to be made from a single-mode device which passes electrons through a series of magnetic-electric barriers built into the device. The barriers assume a wavy spatial profile across the conduction path due to the inevitable broadening of the magnetic fields. Field broadening results in a linearly increasing vector potential across the conduction channel, which increases spin polarization. We have identified that the important factors for generating high spin polarization and conductance modulation are the low source-drain bias, the broadened magnetic fields, and the high number of FM gates within a fixed channel length.

  18. Contemporary surgical devices for male stress urinary incontinence: a review of technological advances in current continence surgery.

    Science.gov (United States)

    Chung, Eric

    2017-07-01

    Male stress urinary incontinence (SUI) remains a debilitating condition that adversely impacts all domains of quality of life and is associated with significant social stigma and health economic burden. The incidence of post-prostatectomy urinary incontinence (PPI) depends on the definition of urinary incontinence and the length of patient follow up. In patients with persistent PPI following failure of conservative measures, surgical treatment is recommended although there is no published guideline on when surgery should be performed, and what the best surgical option is. Male slings (MS) can be divided into adjustable or non-adjustable types, and offers an attractive option for patients who wish to avoid mechanical handling during urinary voiding. Published intermediate data supports good safety and efficacy rate in men with mild to moderate degree of SUI. The AMS 800 artificial urinary sphincter (AUS) remains the standard of treatment for complete continence and has the longest efficacy and safety records. Other AUS-like devices are designed to address current AMS 800 limitations but themselves are fraught with their own issues.

  19. Design of the device of auto-measuring radon continuously based on FPGA

    International Nuclear Information System (INIS)

    Wang Yan; Shen Zhengqin; Chen Qiong

    2004-01-01

    This paper introduces the design of the device of auto-measuring radon continuously. The core of the system is the design of controlling system by FPGA, which consists of preset module, electrical calendar module and driving module. The system can automatically measure the consistence of the radon and the separating out rate of it. The information data is displayed by LCD. The high speed micro printer is used to print the measuring result. It adopts FPGA to design the measuring system of the device, which can improve the precision and stability of the system. (authors)

  20. The Reliability of a Novel Mobile 3-dimensional Wound Measurement Device.

    Science.gov (United States)

    Anghel, Ersilia L; Kumar, Anagha; Bigham, Thomas E; Maselli, Kathryn M; Steinberg, John S; Evans, Karen K; Kim, Paul J; Attinger, Christopher E

    2016-11-01

    Objective assessment of wound dimensions is essential for tracking progression and determining treatment effectiveness. A reliability study was designed to establish intrarater and interrater reliability of a novel mobile 3-dimensional wound measurement (3DWM) device. Forty-five wounds were assessed by 2 raters using a 3DWM device to obtain length, width, area, depth, and volume measurements. Wounds were also measured manually, using a disposable ruler and digital planimetry. The intraclass correlation coefficient (ICC) was used to establish intrarater and interrater reliability. High levels of intrarater and interrater agreement were observed for area, length, and width; ICC = 0.998, 0.977, 0.955 and 0.999, 0.997, 0.995, respectively. Moderate levels of intrarater (ICC = 0.888) and interrater (ICC = 0.696) agreement were observed for volume. Lastly, depth yielded an intrarater ICC of 0.360 and an interrater ICC of 0.649. Measures from the 3DWM device were highly correlated with those obtained from scaled photography for length, width, and area (ρ = 0.997, 0.988, 0.997, P device yielded correlations of ρ = 0.990, 0.987, 0.996 with P device was found to be highly reliable for measuring wound areas for a range of wound sizes and types as compared to manual measurement and digital planimetry. The depth and therefore volume measurement using the 3DWM device was found to have a lower ICC, but volume ICC alone was moderate. Overall, this device offers a mobile option for objective wound measurement in the clinical setting.

  1. Evaluation efficiency for mobile as discharge measurement device for partially filled circular channel

    Directory of Open Access Journals (Sweden)

    I.M.H. Rashwan

    2013-06-01

    Full Text Available Discharge measurement is an important step in engineering applications such as sewerage system, irrigation and drainage system. Mobile flume is a simple and low-cost discharge measuring device constructed of two pieces of pipes, one installed vertically inside the other. The circular mobile flume depends on the concept of developing a contracted zone to have a control section. To evaluate the efficiency of this type of device mathematical and experimental studies are presented. Specific energy, discharge and Froude number equations are used to develop mathematical model. Experimental work is carried out to measure the discharge values corresponding to the specific energy values. The experimental data is used to evaluate the mobile flume as a device to measure discharge. A new equation for the discharge coefficient is presented. This equation can be used to determined accurate discharge with knowing the contraction ratio and measured specific energy.

  2. Development of a Pain Measurement Device Using Electrical Stimulation and Pressure: A Pilot Study

    Directory of Open Access Journals (Sweden)

    Seung Ho Lee

    2018-01-01

    Full Text Available Understanding and precise assessment of pain level are key factors in rehabilitation therapy. Pain is a complex and subjective experience that is affected by an individual’s emotion and health conditions. Various methods have been developed for quantitative evaluation of pain level; however, these methods have several drawbacks. In this work, we developed a pain measurement device for quantitative pain assessment. The system consists of two parts, a component for electrical stimulation and a pressure dolorimeter, for application of two different stresses. Regarding electrical stimulation, the degree of pain is assessed by the applied current. Skin resistance was also analyzed by applying current to remove the effects caused by skin conditions. The electrical stimulation did not induce any histological changes or inflammation in the tissues. Using the pressure dolorimeter, the pain level was assessed according to the degree of inflammation. This system could be used for the quantitative assessment of pain induced by inflammation, wounds, and other factors. Since the described system is the first of its kind, there are many problems that remain to be solved. However, with continuous development, our system could provide more accurate pain assessment by removing skin condition effects and through cross-validation.

  3. Validation of a laser-assisted wound measurement device in a wound healing model.

    Science.gov (United States)

    Constantine, Ryan S; Bills, Jessica D; Lavery, Lawrence A; Davis, Kathryn E

    2016-10-01

    In the treatment and monitoring of a diabetic or chronic wound, accurate and repeatable measurement of the wound provides indispensable data for the patient's medical record. This study aims to measure the accuracy of the laser-assisted wound measurement (LAWM) device against traditional methods in the measurement of area, depth and volume. We measured four 'healing' wounds in a Play-Doh(®) -based model over five subsequent states of wound healing progression in which the model was irregularly filled in to replicate the healing process. We evaluated the LAWM device against traditional methods including digital photograph assessment with National Institutes of Health ImageJ software, measurements of depth with a ruler and weight-to-volume assessment with dental paste. Statistical analyses included analysis of variance (ANOVA) and paired t-tests. We demonstrate that there are significantly different and nearly statistically significant differences between traditional ruler depth measurement and LAWM device measurement, but there are no statistically significant differences in area measurement. Volume measurements were found to be significantly different in two of the wounds. Rate of percentage change was analysed for volume and depth in the wound healing model, and the LAWM device was not significantly different than the traditional measurement technique. While occasionally inaccurate in its absolute measurement, the LAWM device is a useful tool in the clinician's arsenal as it reliably measures rate of percentage change in depth and volume and offers a potentially aseptic alternative to traditional measurement techniques. © 2014 The Authors. International Wound Journal © 2014 Medicalhelplines.com Inc and John Wiley & Sons Ltd.

  4. Current generation by unidirectional lower hybrid waves in the ACT-1 toroidal device

    International Nuclear Information System (INIS)

    Wong, K.L.; Horton, R.; Ono, M.

    1980-05-01

    An unambiguious experimental observation of current generation by unidirectional lower hybrid waves in a toroidal plasma is reported. Up to 10 amperes of current was driven by 500 watts of rf power at 160 MHz

  5. 40 CFR 63.9525 - What are the installation, operation, and maintenance requirements for my weight measurement device?

    Science.gov (United States)

    2010-07-01

    ... measurement device? (a) If you use a solvent recovery system, you must install, operate, and maintain a weight... solvent mixer. If the weight measurement device cannot reproduce the value of the calibration weight..., and maintenance requirements for my weight measurement device? 63.9525 Section 63.9525 Protection of...

  6. Current trends for improving the design of membrane devices for photoautotrophic biosynthesis is light dependent microorganisms

    Directory of Open Access Journals (Sweden)

    A. A. Shevtsov

    2016-01-01

    Full Text Available Modern trends in improving the design of membrane devices for photoautotrophic biosynthesis dependent lighting microorganisms aimed at a significant increase in the productivity of valuable products from biomass of microalgae and obtaining on the basis of their individual useful substances (drugs used in various industries and medicine. In film devices effectively the processes of heat - and mass-exchange with the gas comes into contact with the culture fluid flowing as a film on a transparent film-forming surface is STI in its light intensity and autotrophic biosynthesis occurs only in the presence of a mixture of air with carbon dioxide. Thus, completely eliminated the accumulation of metabolic products due to their continuous removal from film culture liquid with the process gas, which is not typical for devices of other types. Small size membrane bioreactors may increase the degree of saturation of the liquid carbon dioxide with the possibility of changing the concentration of gas in the culture fluid and to ensure the cultivation of microorganisms with a specified biomass yield. At present up to date-developed a significant number of ways to ensure contact of the gas with the liquid (bubbling, gas-lift, mechanical stirring, jet, membrane, etc. on the basis of which an industrial bioreactor, with various "stress" effect. It is believed that for the cultivation of the most optimal are bioreactors with mechanical stirring of the liquid, which allow the greatest productivity of biomass. However, the applied model of a mechanical mixing device to create a work whose cavity of the bioreactor chaotic, disorganized mixing, which contributes to the emergence, insufficient for the sustenance of the cell cultures and microorganisms. Analysis of the interactions of the gas with the liquid film devices showed the need to create a new generation of bioreactor with intensive mass transfer without the possibility of limiting the productivity of

  7. Analytical Expressions for Harmonic Distortion at Low Frequencies due to Device Mismatch in CMOS Current Mirrors

    DEFF Research Database (Denmark)

    Bruun, Erik

    1999-01-01

    One of the origins of harmonic distortion in current mirrors is the inevitable mismatch between the mirror transistors. In this brief we examine both single current mirrors and complementary class AB current mirrors and develop analytical expressions for the mismatch induced harmonic distortion. ...

  8. Measuring Visual Function Using the MultiQuity System: Comparison with an Established Device

    Directory of Open Access Journals (Sweden)

    Patrycja Smolarek-Kasprzak

    2014-01-01

    Full Text Available Purpose. To compare measures of visual acuity (VA and contrast sensitivity (CS from the Thompson Xpert 2000 and MultiQuity (MiQ devices. Methods. Corrected distance VA (CDVA and CS were measured in the right eye of 73 subjects, on an established system (Thompson Xpert and a novel system (MiQ 720. Regression was used to convert MiQ scores into the Thompson scale. Agreement between the converted MiQ and Thompson scores was investigated using standard agreement indices. Test-retest variability for both devices was also investigated, for a separate sample of 24 subjects. Results. For CDVA, agreement was strong between the MiQ and Thomson devices (accuracy = 0.993, precision = 0.889, CCC = 0.883. For CS, agreement was also strong (accuracy = 0.996, precision = 0.911, CCC = 0.907. Agreement was unaffected by demographic variables or by presence/absence of ocular pathology. Test-retest agreement indices for both devices were excellent: in the range 0.88–0.96 for CDVA and in the range 0.90–0.98 for CS. Conclusion. MiQ measurements exhibit strong agreement with corresponding Thomson measurements, and test-retest results are good for both devices. Agreement between the two devices is unaffected by age or ocular pathology.

  9. Zero-field precession and hysteretic threshold currents in a spin torque nano device with tilted polarizer

    Energy Technology Data Exchange (ETDEWEB)

    Zhou Yan; Bonetti, S; Zha, C L; Akerman, Johan [Department of Microelectronics and Applied Physics, Royal Institute of Technology, Electrum 229, 164 40 Kista (Sweden)], E-mail: zhouyan@kth.se

    2009-10-15

    Using nonlinear system theory and numerical simulations, we map out the static and dynamic phase diagrams in the zero applied field of a spin torque nano device with a tilted polarizer (TP). We find that for sufficiently large currents, even very small tilt angles ({beta}>1 deg.) will lead to steady free layer precession in zero field. Within a rather large range of tilt angles, 1 deg. <{beta}<19 deg., we find coexisting static states and hysteretic switching between these using only current. In a more narrow window (1 deg. <{beta}<5 deg.) one of the static states turns into a limit cycle (precession). The coexistence of current-driven static and dynamic states in the zero magnetic field is unique to the TP device and leads to large hysteresis in the upper and lower threshold currents for its operation. The nano device with TP can facilitate the generation of large amplitude mode of spin torque signals without the need for cumbersome magnetic field sources and thus should be very important for future telecommunication applications based on spin transfer torque effects.

  10. Design and development of a device to measure the deformities of clubfoot.

    Science.gov (United States)

    Khas, Kanwaljit S; Pandey, Pulak M; Ray, Alok R

    2015-03-01

    Clubfoot describes a range of foot abnormalities usually present at birth, in which the foot of a baby is twisted out of shape or position. In order to develop an effective treatment plan for clubfoot and/or assess the extent to which existing interventions are successful, medical practitioners need to be able to accurately measure the nature and extent of the deformity. This is typically performed using a goniometer. However, this device is only able to measure one dimension at a time. As such, a complete assessment of the condition of a foot can be extremely burdensome and time-consuming. This article describes a new device that can quickly and efficiently take several measurements on feet of various sizes and shapes. The use of this device was verified by measuring the deformities of real clubfeet. A silicone rubber clubfoot model was also used in this study to clearly illustrate the effectiveness with which the proposed device can measure the various deformities of clubfoot. It is envisaged that the use of this device will significantly reduce the time and effort orthopedists require to measure clubfoot deformities and develop and assess treatment plans. © IMechE 2015.

  11. Transverse current at the periphery of the start plasma of the AMBAL-M open device

    International Nuclear Information System (INIS)

    Taskaev, S.Yu.

    1999-01-01

    Essential longitudinal electron current was experimentally detected earlier in a start plasma of AMBAL-M open trap and system. This paper studies the current at a plasma periphery and the process resulting in occurrence of transverse current, and determines the value of a transverse current in a transporting range. This study enables to determine that filling of a trap with a thin circular flow and formation of nonequilibrium distribution of a radial electrical field in a plasma by potentials of an electrode front gun are the indium peculiarities of the experiment ensuring essential transverse current [ru

  12. Access to hands-on mathematics measurement activities using robots controlled via speech generating devices: three case studies.

    Science.gov (United States)

    Adams, Kim; Cook, Al

    2014-07-01

    To examine how using a robot controlled via a speech generating device (SGD) influences the ways students with physical and communication limitations can demonstrate their knowledge in math measurement activities. Three children with severe physical disabilities and complex communication needs used the robot and SGD system to perform four math measurement lessons in comparing, sorting and ordering objects. The performance of the participants was measured and the process of using the system was described in terms of manipulation and communication events. Stakeholder opinions were solicited regarding robot use. Robot use revealed some gaps in the procedural knowledge of the participants. Access to both the robot and SGD was shown to provide several benefits. Stakeholders thought the intervention was important and feasible for a classroom environment. The participants were able to participate actively in the hands-on and communicative measurement activities and thus meet the demands of current math instruction methods. Current mathematics pedagogy encourages doing hands-on activities while communicating about concepts. Adapted Lego robots enabled children with severe physical disabilities to perform hands-on length measurement activities. Controlling the robots from speech generating devices (SGD) enabled the children, who also had complex communication needs, to reflect and report on results during the activities. By using the robots combined with SGDs, children both exhibited their knowledge of and experienced the concepts of mathematical measurements.

  13. An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

    International Nuclear Information System (INIS)

    Polee, C; Chankow, N; Srisatit, S; Thong-Aram, D

    2015-01-01

    In film radiography, underexposure and overexposure may happen particularly when lacking information of specimen material and hollowness. This paper describes a method and a device for determining exposure in industrial gamma-ray radiography based on quick measurement of transmitted gamma-ray intensity with a small detector. Application software was developed for Android mobile phone to remotely control the device and to display counting data via Bluetooth communication. Prior to film exposure, the device is placed behind a specimen to measure transmitted intensity which is inversely proportional to the exposure. Unlike in using the conventional exposure curve, correction factors for source decay, source-to- film distance, specimen thickness and kind of material are not needed. The developed technique and device make radiographic process economic, convenient and more reliable. (paper)

  14. Implantable optogenetic device with CMOS IC technology for simultaneous optical measurement and stimulation

    Science.gov (United States)

    Haruta, Makito; Kamiyama, Naoya; Nakajima, Shun; Motoyama, Mayumi; Kawahara, Mamiko; Ohta, Yasumi; Yamasaki, Atsushi; Takehara, Hiroaki; Noda, Toshihiko; Sasagawa, Kiyotaka; Ishikawa, Yasuyuki; Tokuda, Takashi; Hashimoto, Hitoshi; Ohta, Jun

    2017-05-01

    In this study, we have developed an implantable optogenetic device that can measure and stimulate neurons by an optical method based on CMOS IC technology. The device consist of a blue LED array for optically patterned stimulation, a CMOS image sensor for acquiring brain surface image, and eight green LEDs surrounding the CMOS image sensor for illumination. The blue LED array is placed on the CMOS image sensor. We implanted the device in the brain of a genetically modified mouse and successfully demonstrated the stimulation of neurons optically and simultaneously acquire intrinsic optical images of the brain surface using the image sensor. The integrated device can be used for simultaneously measuring and controlling neuronal activities in a living animal, which is important for the artificial control of brain functions.

  15. The Measurement of Spectral Characteristics and Composition of Radiation in Atlas with MEDIPIX2-USB Devices

    Science.gov (United States)

    Campbell, M.; Doležal, Z.; Greiffenberg, D.; Heijne, E.; Holy, T.; Idárraga, J.; Jakůbek, J.; Král, V.; Králík, M.; Lebel, C.; Leroy, C.; Llopart, X.; Lord, G.; Maneuski, D.; Ouellette, O.; Sochor, V.; Pospíšil, S.; Suk, M.; Tlustos, L.; Vykydal, Z.; Wilhelm, I.

    2008-06-01

    A network of devices to perform real-time measurements of the spectral characteristics and composition of radiation in the ATLAS detector and cavern during its operation is being built. This system of detectors will be a stand alone system fully capable of delivering real-time images of fluxes and spectral composition of different particle species including slow and fast neutrons. The devices are based on MEDIPIX2 pixel silicon detectors that will be operated via active USB cables and USB-Ethernet extenders through an Ethernet network by a PC located in the USA15 ATLAS control room. The installation of 14 devices inside ATLAS (detector and cavern) is in progress.

  16. The Measurement of Spectral Characteristics and Composition of Radiation in ATLAS with MEDIPIX2-USB Devices

    CERN Document Server

    Campbell, M.; Greiffenberg, D.; Heijne, E.; Holy, T.; Idárraga, J.; Jakubek, J.; Král, V.; Králík, M.; Lebel, C.; Leroy, C.; Llopart, X.; Lord, G.; Maneuski, D.; Ouellette, O.; Sochor, V.; Prospísil, S.; Suk, M; Tlustos, L.; Vykydal, Z.; Wilhelm, I.

    2008-01-01

    A network of devices to perform real-time measurements of the spectral characteristics and composition of radiation in the ATLAS detector and cavern during its operation is being built. This system of detectors will be a stand alone system fully capable of delivering real-time images of fluxes and spectral composition of different particle species including slow and fast neutrons. The devices are based on MEDIPIX2 pixel silicon detectors that will be operated via active USB cables and USB-Ethernet extenders through an Ethernet network by a PC located in the USA15 ATLAS control room. The installation of 14 devices inside ATLAS (detector and cavern) is in progress.

  17. [Magnetic resonance imaging in patients with implantable devices for treatment of disturbed heart rhythm: review of the current situation].

    Science.gov (United States)

    Sviridova, A A

    The question of the possibility of MRI scanning in patients with cardiac implantable electronic devices (CIED) appeared simultaneously with the introduction of MRI in clinical practice. A lot of in-vitro, in-vivo and clinical researches were performed to estimate wat going on with CIED in strong magnetic field and is it possible to perform some unified protocol of safe MRI-scanning for these patients. Recommendations were provided, but not for the wide practice. MRI remained strongly contraindicated for CIED patient. To meet the clinical need CIEM manufacturers changed the design of devices to made them MRI-compatible, including reducing of ferromagnetic components, additional filters, new software. Lead coil design was changed as well to minimize lead heating and electrical current induction. Now all leaders of CIED industry have in their portfolio all types of MRI-conditional implanted cardiac rhythm management devices (pacemakers, ICDs, CRTs). "Conditional" means MRI scanning can be done only under specific condition. For MRI device and lead in one system have to be from the same manufacturer. Now, if you need to implant the device, you must proceed from the fact that the patient is more likely to need an MRI in the future and choose the appropriate model, not forgetting that the electrodes should also be MRI-compatible.

  18. Compact and high-efficiency device for Raman scattering measurement using optical fibers.

    Science.gov (United States)

    Mitsui, Tadashi

    2014-11-01

    We describe the design and development of a high-efficiency optical measurement device for operation within the small bore of a high-power magnet at low temperature. For the high-efficiency measurement of light emitted from this small region, we designed a compact confocal optics with lens focusing and tilting systems, and used a piezodriven translation stage that allows micron-scale focus control of the sample position. We designed a measurement device that uses 10 m-long optical fibers in order to avoid the influence of mechanical vibration and magnetic field leakage of high-power magnets, and we also describe a technique for minimizing the fluorescence signal of optical fibers. The operation of the device was confirmed by Raman scattering measurements of monolayer graphene on quartz glass with a high signal-to-noise ratio.

  19. Measuring device and method for dimples height differences of 17 x 17 grid

    International Nuclear Information System (INIS)

    Xu Yilan; Zheng Zhihui; Yan Liwei; Wang Xihe

    2001-01-01

    There are 264 cell for fastening fuel rods in the grid of 17 x 17 fuel assembly of PWR. The height differences of top and bottom dimples in a grid is an important quality characteristic of the grid. The report deals with measuring machine and method for dimples height differences of the grid. The device has two measuring probes. The Parallel Leaf Spring is used for transmitting the little displacement between two probes. The uncertainty of the device is σ≤4 μm. The measuring method is shown to be practicable

  20. Plasma electron density measurement with multichannel microwave interferometer on the HL-1 tokamak device

    International Nuclear Information System (INIS)

    Xu Deming; Zhang Hongyin; Liu Zetian; Ding Xuantong; Li Qirui; Wen Yangxi

    1989-11-01

    A multichannel microwave interferometer which is composed of different microwave interferometers (one 2 mm band, one 4 mm band and two 8 mm band) has been used to measure the plasma electron density on HL-1 tokamak device. The electron density approaching to 5 x 10 13 cm -3 is measured by a 2 mm band microwave interferometer. In the determinable range, the electron density profile in the cross-section on HL-1 device has been measured by this interferometer. A microcomputer data processing system is also developed