WorldWideScience

Sample records for calibration transfer procedures

  1. Development of the calibration transfer procedures for the XRF analyser AF-20

    CERN Document Server

    Kowalska, E

    2001-01-01

    Two standardization techniques; Direct (DS) and Picewise Direct Standardization (PDS) were applied to multivariate models developed using x-ray spectra. The data were obtained for the determination of Ca and fe in flying ash from brown coal using low resolution XRF analysis. The analyser AF-20 was used with an argon proportional counter and a Cd-109 radioisotope source. Number and comparison of samples from the calibration set were selected. Root mean square error of prediction (RMSEPr) was used as a criterion for assessment of the standardization quality. The best performance was obtained for the PDS method using 5 samples chosen with the leverage method. The RMSEPr value was 4 times lower for Ca determination and 2 times lower for Fe compared with the results without standardization. Software package was designed for fast computation of the standardization transform. The program can be used when the instrument parameters have changed as well as for calibration of the new instrument.

  2. Instrument Calibration and Certification Procedure

    Energy Technology Data Exchange (ETDEWEB)

    Davis, R. Wesley [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2016-05-31

    The Amptec 640SL-2 is a 4-wire Kelvin failsafe resistance meter, designed to reliably use very low-test currents for its resistance measurements. The 640SL-1 is a 2-wire version, designed to support customers using the Reynolds Industries type 311 connector. For both versions, a passive (analog) dual function DC Milliameter/Voltmeter allows the user to verify the actual 640SL output current level and the open circuit voltage on the test leads. This procedure includes tests of essential performance parameters. Any malfunction noticed during calibration, whether specifically tested for or not, shall be corrected before calibration continues or is completed.

  3. Flexible calibration procedure for fringe projection profilometry

    OpenAIRE

    Vargas, Javier; Quiroga Mellado, Juan Antonio; Terrón López, María José

    2007-01-01

    A novel calibration method for whole field three-dimensional shape measurement by means of fringe projection is presented. Standard calibration techniques, polynomial-and model-based, have practical limitations such as the difficulty of measuring large fields of view, the need to use precise z stages, and bad calibration results due to inaccurate calibration points. The proposed calibration procedure is a mixture of the two main standard techniques, sharing their benefits and avoiding their m...

  4. Calibration procedure for fire resistance furnaces

    NARCIS (Netherlands)

    Twilt, L.; Leur, P.H.E. van de; Wickström, U.

    1996-01-01

    On behalf of CEN/TC 127 "Fire Safety in Buildings", a series of tests has been carried out to evaluate and complete the draft calibration procedure for fire resistance furnaces [4]. Fourteen laboratories in nine European countries participated in the test series, each carrying out one calibration te

  5. Calibration of routine dosimeters in radiation processing: Validation procedure for in-plant calibration

    Directory of Open Access Journals (Sweden)

    Šećerov Bojana Lj.

    2011-01-01

    Full Text Available The essential prerequisite of radiation dosimetry is to provide quality assurance and documentation that the irradiation procedure has been carried out according to the specification requirement of correct calibration of the chosen dosimetry system. At the Radiation Plant of the Vinča Institute of Nuclear Sciences we compared two recommended protocols of irradiation procedures in the calibration of dosimetry systems in radiation processing: (1 by irradiation of routine dosimeters (ethanol-chlorobenzene - ECB at the calibration laboratory and (2, by in-plant calibration with alanine transfer - dosimeters. The critical point for in-plant calibration is irradiation geometry, so we carefully positioned the phantom carrying both dosimeters in order to minimize dose gradients across the sample. The analysis of results obtained showed that the difference among determined absorbed doses for the construction of calibration curves between these two methods, (alanine vs. ECB, is less than 1%. The difference in combined standard uncertainty for each calibration procedure is 0.1%. These results demonstrate that our in-plant calibration is as good as calibration by irradiation at the calibration laboratory and validates our placement of the irradiation phantom during irradiation.

  6. AFFTC Standard Airspeed Calibration Procedures

    Science.gov (United States)

    1981-06-01

    25x0UIXQXQ Results of groundLpeed course calibration are normally pre- sented in the following plots: 1. .AvP vs Vi Ŗ. All vs V ic 3. AMPC vs Mic .4...8217Average AfPeavgpo, tion correction AM /AH 10-5 per and figure V 9 PC PC feet . fu V AYpc" x q3 @ , Average position avg corred ion (AM @ AMPC /AVPC...instrument error 0 M ic From and 0), Mach number p Chart 8.5 in reference’l (AFTR 6273) (DO AMPPacer poqition error calibra- Pc tion at9 S( AMpc /’,HpC)p

  7. Calibration Procedures on Oblique Camera Setups

    Science.gov (United States)

    Kemper, G.; Melykuti, B.; Yu, C.

    2016-06-01

    Beside the creation of virtual animated 3D City models, analysis for homeland security and city planning, the accurately determination of geometric features out of oblique imagery is an important task today. Due to the huge number of single images the reduction of control points force to make use of direct referencing devices. This causes a precise camera-calibration and additional adjustment procedures. This paper aims to show the workflow of the various calibration steps and will present examples of the calibration flight with the final 3D City model. In difference to most other software, the oblique cameras are used not as co-registered sensors in relation to the nadir one, all camera images enter the AT process as single pre-oriented data. This enables a better post calibration in order to detect variations in the single camera calibration and other mechanical effects. The shown sensor (Oblique Imager) is based o 5 Phase One cameras were the nadir one has 80 MPIX equipped with a 50 mm lens while the oblique ones capture images with 50 MPix using 80 mm lenses. The cameras are mounted robust inside a housing to protect this against physical and thermal deformations. The sensor head hosts also an IMU which is connected to a POS AV GNSS Receiver. The sensor is stabilized by a gyro-mount which creates floating Antenna -IMU lever arms. They had to be registered together with the Raw GNSS-IMU Data. The camera calibration procedure was performed based on a special calibration flight with 351 shoots of all 5 cameras and registered the GPS/IMU data. This specific mission was designed in two different altitudes with additional cross lines on each flying heights. The five images from each exposure positions have no overlaps but in the block there are many overlaps resulting in up to 200 measurements per points. On each photo there were in average 110 well distributed measured points which is a satisfying number for the camera calibration. In a first step with the help of

  8. CALIBRATION PROCEDURES ON OBLIQUE CAMERA SETUPS

    Directory of Open Access Journals (Sweden)

    G. Kemper

    2016-06-01

    Full Text Available Beside the creation of virtual animated 3D City models, analysis for homeland security and city planning, the accurately determination of geometric features out of oblique imagery is an important task today. Due to the huge number of single images the reduction of control points force to make use of direct referencing devices. This causes a precise camera-calibration and additional adjustment procedures. This paper aims to show the workflow of the various calibration steps and will present examples of the calibration flight with the final 3D City model. In difference to most other software, the oblique cameras are used not as co-registered sensors in relation to the nadir one, all camera images enter the AT process as single pre-oriented data. This enables a better post calibration in order to detect variations in the single camera calibration and other mechanical effects. The shown sensor (Oblique Imager is based o 5 Phase One cameras were the nadir one has 80 MPIX equipped with a 50 mm lens while the oblique ones capture images with 50 MPix using 80 mm lenses. The cameras are mounted robust inside a housing to protect this against physical and thermal deformations. The sensor head hosts also an IMU which is connected to a POS AV GNSS Receiver. The sensor is stabilized by a gyro-mount which creates floating Antenna –IMU lever arms. They had to be registered together with the Raw GNSS-IMU Data. The camera calibration procedure was performed based on a special calibration flight with 351 shoots of all 5 cameras and registered the GPS/IMU data. This specific mission was designed in two different altitudes with additional cross lines on each flying heights. The five images from each exposure positions have no overlaps but in the block there are many overlaps resulting in up to 200 measurements per points. On each photo there were in average 110 well distributed measured points which is a satisfying number for the camera calibration. In a first

  9. NASA Glenn Icing Research Tunnel: 2012 Cloud Calibration Procedure and Results

    Science.gov (United States)

    VanZante, Judith Foss; Ide, Robert F.; Steen, Laura E.

    2012-01-01

    In 2011, NASA Glenn s Icing Research Tunnel underwent a major modification to it s refrigeration plant and heat exchanger. This paper presents the results of the subsequent full cloud calibration. Details of the calibration procedure and results are presented herein. The steps include developing a nozzle transfer map, establishing a uniform cloud, conducting a drop sizing calibration and finally a liquid water content calibration. The goal of the calibration is to develop a uniform cloud, and to build a transfer map from the inputs of air speed, spray bar atomizing air pressure and water pressure to the output of median volumetric droplet diameter and liquid water content.

  10. Calibration Procedure for 3D Turning Dynamometer

    DEFF Research Database (Denmark)

    Axinte, Dragos Aurelian; Belluci, Walter

    1999-01-01

    The aim of the static calibration of the dynamometer is to obtain the matrix for evaluating cutting forces through the output voltage of the piezoelectric cells and charge amplifiers. In the same time, it is worth to evaluate the linearity of the dependencies between applied forces and output...... of the piezoelectric cells;5. Mounting of the dynamometer;6. Calibration of the dynamometer;7. Data analysis;8. Uncertainty budget of the calibration....

  11. Brookhaven National Laboratory meteorological services instrument calibration plan and procedures

    Energy Technology Data Exchange (ETDEWEB)

    Heiser .

    2013-02-16

    This document describes the Meteorological Services (Met Services) Calibration and Maintenance Schedule and Procedures, The purpose is to establish the frequency and mechanism for the calibration and maintenance of the network of meteorological instrumentation operated by Met Services. The goal is to maintain the network in a manner that will result in accurate, precise and reliable readings from the instrumentation.

  12. Calibration procedure for a laser triangulation scanner with uncertainty evaluation

    Science.gov (United States)

    Genta, Gianfranco; Minetola, Paolo; Barbato, Giulio

    2016-11-01

    Most of low cost 3D scanning devices that are nowadays available on the market are sold without a user calibration procedure to correct measurement errors related to changes in environmental conditions. In addition, there is no specific international standard defining a procedure to check the performance of a 3D scanner along time. This paper aims at detailing a thorough methodology to calibrate a 3D scanner and assess its measurement uncertainty. The proposed procedure is based on the use of a reference ball plate and applied to a triangulation laser scanner. Experimental results show that the metrological performance of the instrument can be greatly improved by the application of the calibration procedure that corrects systematic errors and reduces the device's measurement uncertainty.

  13. [Research on calibration transfer of NIR filter spectrophotometer].

    Science.gov (United States)

    Chen, Jia-wei; Zhou, Chang-le; Zhang, Ye-hui; Xu, Xiao-jie; Lin, Kun-hui; Ye, Nan

    2008-10-01

    Calibration transfer is an important issue to building up universal and comparable performance of spectrometer data in near infrared spectral analysis technology. Methods of slope/bias correction, direct standardization (DS), and target factor analysis (TFA) were used for the calibration transfer among five NIR filter spectrophotometers using maize as the samples. The effects of three calibration transfer methods were compared. The DS method has the best performance. The average calibration transfer difference of DS is 7.01%. This study also relates to the dependence of calibration transfer on the number of standardization samples. It was proven by experiment that the results of calibration transfer will be better as the number of samples is increased and will be generally stable when there are twenty standardization samples. However, the effect of calibration transfer attained by DS is degraded sharply when the number of standardization samples decreases to be below twenty. Moreover, slope/bias and TFA are not sensitive to the number of standardization samples.

  14. Generic calibration procedures for nacelle-based profiling lidars

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael; Wagner, Rozenn

    In power performance testing, it has been demonstrated that the effects of wind speed and direction variations over the rotor disk can no longer be neglected for large wind turbines [1]. A new generation of commercial nacelle-based lidars is now available, offering wind profiling capabilities...... to calibrate profiling nacelle lidars........ Developing standard procedures for power curves using lidars requires assessing lidars measurement uncertainty that is provided by a calibration. Based on the calibration results from two lidars, the Avent 5-beam Demonstrator and the Zephir Dual Mode (ZDM), we present in this paper a generic methodology...

  15. Simple transfer calibration method for a Cimel Sun-Moon photometer: calculating lunar calibration coefficients from Sun calibration constants.

    Science.gov (United States)

    Li, Zhengqiang; Li, Kaitao; Li, Donghui; Yang, Jiuchun; Xu, Hua; Goloub, Philippe; Victori, Stephane

    2016-09-20

    The Cimel new technologies allow both daytime and nighttime aerosol optical depth (AOD) measurements. Although the daytime AOD calibration protocols are well established, accurate and simple nighttime calibration is still a challenging task. Standard lunar-Langley and intercomparison calibration methods both require specific conditions in terms of atmospheric stability and site condition. Additionally, the lunar irradiance model also has some known limits on its uncertainty. This paper presents a simple calibration method that transfers the direct-Sun calibration constant, V0,Sun, to the lunar irradiance calibration coefficient, CMoon. Our approach is a pure calculation method, independent of site limits, e.g., Moon phase. The method is also not affected by the lunar irradiance model limitations, which is the largest error source of traditional calibration methods. Besides, this new transfer calibration approach is easy to use in the field since CMoon can be obtained directly once V0,Sun is known. Error analysis suggests that the average uncertainty of CMoon over the 440-1640 nm bands obtained with the transfer method is 2.4%-2.8%, depending on the V0,Sun approach (Langley or intercomparison), which is comparable with that of lunar-Langley approach, theoretically. In this paper, the Sun-Moon transfer and the Langley methods are compared based on site measurements in Beijing, and the day-night measurement continuity and performance are analyzed.

  16. A multivariate calibration procedure for the tensammetric determination of detergents

    NARCIS (Netherlands)

    Bos, M.

    1989-01-01

    A multivariate calibration procedure based on singular value decomposition (SVD) and the Ho-Kashyap algorithm is used for the tensammetric determination of the cationic detergents Hyamine 1622, benzalkonium chloride (BACl), N-cetyl-N,N,N-trimethylammonium bromide (CTABr) and mixtures of CTABr and BA

  17. [Calibration transfer between two FTNIR spectrophotometers using SVR].

    Science.gov (United States)

    Zhao, Long-lian; Li, Jun-hui; Zhang, Wen-juan; Wang, Jian-cai; Zhang, Lu-da

    2008-10-01

    In the present research, a set of maize powder samples was used to study the calibration transfer between two fourier transform near-infrared (FTNIR) spectrophotometers, and a method of moving window support vector regression machines (SVR) was used to correct the differences between the two instruments. Bruker Vector 22/N was referred to as "master" on which the maize protein calibration model was built. Bruker MPA was referred to as "slave" instrument. A transformation matrix was constructed based on the spectra of a sample set (for calibration transfer) measured on both instruments. After transfer, NIR spectra acquired on "slave" will appear as if they were measured on master instrument. The calibration model available for the master can then be used to predict the transformed spectra measured on the slave. The transfer parameters were computed as follows. For wavelength i, the absorbance vector obtained on the master instrument was regressed against the corresponding absorbance matrix of a spectral window obtained on the slave instrument. Method of SVR was used for regression Moving the wavelength i and corresponding window, the transfer parameter for each wavelength can be obtained. For the two FTNIR spectrophotometers, a window size of 31 wavelengths and a subset of 15 transfer samples were chosen to establish the SVR regression model between "master" and "slave". Applying the calibration model to the prediction samples after being corrected by the transfer parameters, a good transfer performance can be achieved. The correlation coefficient (r) is 0.9434, while the relative standard deviation (RSD) is 4.23%. These results suggest that the SVR method can be used to successfully transfer the calibration model for protein of maize developed on a FTNIR spectrophotometer to another.

  18. Calibration procedure of measuring system for vehicle wheel load estimation

    Science.gov (United States)

    Kluziewicz, M.; Maniowski, M.

    2016-09-01

    The calibration procedure of wheel load measuring system is presented. Designed method allows estimation of selected wheel load components while the vehicle is in motion. Mentioned system is developed to determine friction forces between tire and road surface, basing on measured internal reaction forces in wheel suspension mechanism. Three strain gauge bridges and three-component piezoelectric load cell are responsible for internal force measurement in suspension components, two wire sensors are measuring displacements. External load is calculated via kinematic model of suspension mechanism implemented in Matlab environment. In the described calibration procedure, internal reactions are measured on a test stand while the system is loaded by a force of known direction and value.

  19. 16 CFR 1209.8 - Procedure for calibration of radiation instrumentation.

    Science.gov (United States)

    2010-01-01

    ... 16 Commercial Practices 2 2010-01-01 2010-01-01 false Procedure for calibration of radiation instrumentation. 1209.8 Section 1209.8 Commercial Practices CONSUMER PRODUCT SAFETY COMMISSION CONSUMER PRODUCT... for calibration of radiation instrumentation. This procedure is used to calibrate the...

  20. Engineering task plan for rotary mode core sample truck calibration procedure support engineering

    Energy Technology Data Exchange (ETDEWEB)

    BOGER, R.M.

    1999-06-04

    This Engineering Task Plan defines the tasks associated with providing an overview, with respect to the Authorization Basis and PHMC procedures, as applicable, of the calibration documentation for operational equipment for which Characterization Engineering has calibration responsibility.

  1. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Science.gov (United States)

    2010-10-01

    ..., if possible, traceable to a reference method or reference material of known value; and (ii) Including... 42 Public Health 5 2010-10-01 2010-10-01 false Standard: Calibration and calibration verification... for Nonwaived Testing Analytic Systems § 493.1255 Standard: Calibration and calibration...

  2. 40 CFR 89.422 - Dilute sampling procedures-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... NIST measurements, and will serve as the reference value (NIST “true” value) for the CVS calibration... calibration curve has been obtained, verification of the entire system may be performed by injecting a known... verification procedure is found in paragraph (e) of this section. (c) PDP-CVS calibration. (1) The...

  3. A least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer.

    Science.gov (United States)

    Ribeiro, F B; Carlos, D U; Hiodo, F Y; Strobino, E F

    2005-01-01

    In this study, a least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer using the general inverse matrix method is presented. The procedure weights the model equations fitting to the calibration data, taking into account the variances in the counting rates and in the radioactive standard concentrations. The application of the described procedure is illustrated by calibrating twice the same gamma-ray spectrometer, with two independent data sets collected approximately 18 months apart in the same calibration facility.

  4. An improved outdoor calibration procedure for broadband ultraviolet radiometers.

    Science.gov (United States)

    Cancillo, M L; Serrano, A; Antón, M; García, J A; Vilaplana, J M; de la Morena, B

    2005-01-01

    This article aims at improving the broadband ultraviolet radiometer's calibration methodology. For this goal, three broadband radiometers are calibrated against a spectrophotometer of reference. Three different one-step calibration models are tested: ratio, first order and second order. The latter is proposed in order to adequately reproduce the high dependence on the solar zenith angle shown by the other two models and, therefore, to improve the calibration performance at high solar elevations. The proposed new second-order model requires no additional information and, thus, keeps the operational character of the one-step methodology. The models are compared in terms of their root mean square error and the most qualified is subsequently validated by comparing its predictions with the spectrophotometer measurements within an independent validation data subset. Results show that the best calibration is achieved by the second-order model, with a mean bias error and mean absolute bias error lower than 2.2 and 6.7%, respectively.

  5. Standard Procedure for Calibrating an Areal Calorimetry Based Dosimeter

    Science.gov (United States)

    2015-05-01

    unlimited (P.A. Case No. TSRL-PA-2015-0136, 24 Sep 15). REFERENCES Beason, C. W. (2005). CLT As A Calorimeter. PowerPoint presentation dated 10 January...or sell any patented invention that may relate to them. This report was cleared for public release by the 88th ABW Public Affairs Office and is...is presented for calibrating an areal calorimetry-based dosimeter. A detailed example of calibrating an areal calorimetry-based dosimeter is

  6. Calibration procedure for 3D surface measurements using stereo vision and laser stripe

    OpenAIRE

    Vilaça, João L.; Fonseca, Jaime C.; Pinho, A. C. Marques de

    2006-01-01

    This paper proposes a new stereo vision calibration procedure and laser strip detection for 3D surface measurements. In this calibration procedure the laser plane is the one that matters, only one set of laser-coplanar calibration points is needed for image cameras calibration; and a dead- zone scan area is considered, since the igitalization arm is assembled in a 3 degree-freedom machine PC-based Motion Control with multiple scan paths. It is also presented some algorithms for 3D surface tre...

  7. Resolution of a phase ambiguity in a calibration procedure for polarimetric radar systems

    Energy Technology Data Exchange (ETDEWEB)

    Sletten, M.A. (Naval Research Lab., Washington, DC (United States). Radar Div.)

    1994-01-01

    In response to the remote sensing communities' interest in radar polarimetry, considerable effort has recently been devoted to the development of calibration techniques for polarimetric radar systems. A cross-pol/co-pol phase ambiguity in a previously published calibration procedure for polarimetric radar systems is discussed. The original procedure is modified to resolve the ambiguity while still retaining insensitivity to calibration target orientation. The modified form is then generalized and applied to an ultrawideband radar system for which the ambiguity in the original procedure is particularly evident.

  8. 33 CFR 155.750 - Contents of transfer procedures.

    Science.gov (United States)

    2010-07-01

    ...) Exemptions or alternatives granted must be placed in the front of the transfer procedures. (c) The vessel... accordance with 46 CFR 39.30-1(d) (1) through (d)(3); (4) The initial transfer rate for each tank that complies with 46 CFR 39.30-1(h); (5) A table or graph of transfer rates and corresponding vapor...

  9. 40 CFR 91.424 - Dilute sampling procedure-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... for Standards and Testing (NIST), and will serve as the reference value (NIST “true” value) for the... approved in advance by the Administrator. (b) After the calibration curve has been obtained, verification...., analyzer calibration, leaks, or HC hangup). A verification procedure is found in paragraph (e) of...

  10. NASA Glenn Icing Research Tunnel: 2014 Cloud Calibration Procedure and Results

    Science.gov (United States)

    Van Zante, Judith F.; Ide, Robert F.; Steen, Laura E.; Acosta, Waldo J.

    2014-01-01

    The results of the December 2013 to February 2014 Icing Research Tunnel full icing cloud calibration are presented. The calibration steps included establishing a uniform cloud and conducting drop size and liquid water content calibrations. The goal of the calibration was to develop a uniform cloud, and to generate a transfer function from the inputs of air speed, spray bar atomizing air pressure and water pressure to the outputs of median volumetric drop diameter and liquid water content. This was done for both 14 CFR Parts 25 and 29, Appendix C ('typical' icing) and soon-to-be released Appendix O (supercooled large drop) conditions.

  11. Time Transfer Through Optical Fibers (TTTOF): First Results of Calibrated Clock Comparisons

    CERN Document Server

    Piester, D; Rost, M; Bauch, A

    2010-01-01

    We have developed a means for accurate time transfer using optical fibers and aim at the synchronization of clocks located at different places on an institute campus with an overall uncertainty of 100 ps or better. Such an installation shall be used as a part of the infrastructure connecting the ground station setups during forthcoming T2L2 and ACES experiments and the local installations at the PTB time laboratory. Our target transmission length is less than 1 km. To transfer time, a code-domain-multiple-access (CDMA) signal is used for modulation of a laser. Optical signals are exchanged in the two-way mode to cancel long-term fiber length variation. This is similar to the well known two-way satellite time and frequency transfer (TWSTFT) scheme. We discuss procedures for a proper calibration of such time transfer through optical fibers links and show first promising results of an experiment using a test loop on the PTB campus with a length of 2 km.

  12. Calibrated Pulse-Thermography Procedure for Inspecting HDPE

    Directory of Open Access Journals (Sweden)

    Mohammed A. Omar

    2008-01-01

    Full Text Available This manuscript discusses the application of a pulse-thermography modality to evaluate the integrity of a high-density polyethylene HDPE joint for delamination, in nonintrusive manner. The inspected HDPE structure is a twin-cup shape, molded through extrusion, and the inspection system comprises a high-intensity, short-duration radiation pulse to excite thermal emission; the text calibrates the experiment settings (pulse duration, and detector sampling rate to accommodate HDPE bulks thermal response. The acquired thermal scans are processed through new contrast computation named “self-referencing”, to investigate the joint tensile strength and further map its adhesion interface in real-time. The proposed system (hardware, software combination performance is assessed through an ultrasound C-scan validation and further benchmarked using a standard pulse phase thermography (PPT routine.

  13. Modified calibration procedures for a Yankee Environmental System UVB-1 biometer based on spectral measurements with a brewer spectrophotometer.

    Science.gov (United States)

    Vilaplana, José M; Cachorro, Victoria E; Sorribas, Mar; Luccini, Eduardo; de Frutos, Angel M; Berjón, Alberto; de la Morena, Benito

    2006-01-01

    The calibration of the erythemal irradiance measured by a Yankee Environmental System (YES) UVB-1 biometer is presented using two methods of calibration with a wide range of experimental solar zenith angles (SZAs) and ozone values. The calibration is performed through simultaneous spectral measurements by a calibrated double-monochromator Brewer MK-III spectrophotometer at "El Arenosillo" station, located in southwestern Spain. Because the range of spectral measurements of the Brewer spectrophotometer is 290-363 nm, a previously validated radiative transfer model was used to account for the erythemal contribution between 363 and 400 nm. Both methods are recommended by the World Meteorological Organization and we present and discuss here a wide range of results and features given by modified procedures applied to these two general methods. As is well established, the calibration factor for this type of radiometric system is dependent on atmospheric conditions, the most important of which are the ozone content and the SZA. Although the first method is insensitive to these two factors, we analyze this behavior in terms of the range used for the SZA and the use of two different mathematical approaches for its determination. The second method shows the dependence on SZA and ozone content and, thus, a polynomial as a function of SZA or a matrix including SZA and ozone content were determined as general calibration factors for the UV radiometric system. We must note that the angular responses of the YES radiometer and Brewer spectroradiometer have not been considered, because of the difficulty in correcting them. The results show in detail the advantages and drawbacks (and the corresponding associated error) given by the different approaches used for the determination of these calibration coefficients.

  14. A recommendation for revised dose calibrator measurement procedures for 89Zr and 124I.

    Directory of Open Access Journals (Sweden)

    Bradley J Beattie

    Full Text Available Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R using a published calibration setting number of 465. PET images of (124I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89Zr and (124I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an 18F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89Zr and (124I. These and other positron emitting radionuclides can be accurately calibrated relative to 18F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances.

  15. Mathematical calibration procedure of a capacitive sensor-based indexed metrology platform

    Science.gov (United States)

    Brau-Avila, A.; Santolaria, J.; Acero, R.; Valenzuela-Galvan, M.; Herrera-Jimenez, V. M.; Aguilar, J. J.

    2017-03-01

    The demand for faster and more reliable measuring tasks for the control and quality assurance of modern production systems has created new challenges for the field of coordinate metrology. Thus, the search for new solutions in coordinate metrology systems and the need for the development of existing ones still persists. One example of such a system is the portable coordinate measuring machine (PCMM), the use of which in industry has considerably increased in recent years, mostly due to its flexibility for accomplishing in-line measuring tasks as well as its reduced cost and operational advantages compared to traditional coordinate measuring machines. Nevertheless, PCMMs have a significant drawback derived from the techniques applied in the verification and optimization procedures of their kinematic parameters. These techniques are based on the capture of data with the measuring instrument from a calibrated gauge object, fixed successively in various positions so that most of the instrument measuring volume is covered, which results in time-consuming, tedious and expensive verification and optimization procedures. In this work the mathematical calibration procedure of a capacitive sensor-based indexed metrology platform (IMP) is presented. This calibration procedure is based on the readings and geometric features of six capacitive sensors and their targets with nanometer resolution. The final goal of the IMP calibration procedure is to optimize the geometric features of the capacitive sensors and their targets in order to use the optimized data in the verification procedures of PCMMs.

  16. Radiation Heat Transfer Procedures for Space-Related Applications

    Science.gov (United States)

    Chai, John C.

    2000-01-01

    Over the last contract year, a numerical procedure for combined conduction-radiation heat transfer using unstructured grids has been developed. As a result of this research, one paper has been published in the Numerical Heat Transfer Journal. One paper has been accepted for presentation at the International Center for Heat and Mass Transfer's International Symposium on Computational Heat Transfer to be held in Australia next year. A journal paper is under review by my NASA's contact. A conference paper for the ASME National Heat Transfer conference is under preparation. In summary, a total of four (4) papers (two journal and two conference) have been published, accepted or are under preparation. There are two (2) to three (3) more papers to be written for the project. In addition to the above publications, one book chapter, one journal paper and six conference papers have been published as a result of this project. Over the last contract year, the research project resulted in one Ph.D. thesis and partially supported another Ph.D. student. My NASA contact and myself have formulated radiation heat transfer procedures for materials with different indices of refraction and for combined conduction-radiation heat transfer. We are trying to find other applications for the procedures developed under this grant.

  17. Design of an ultra-portable field transfer radiometer supporting automated vicarious calibration

    Science.gov (United States)

    Anderson, Nikolaus; Thome, Kurtis; Czapla-Myers, Jeffrey; Biggar, Stuart

    2015-09-01

    The University of Arizona Remote Sensing Group (RSG) began outfitting the radiometric calibration test site (RadCaTS) at Railroad Valley Nevada in 2004 for automated vicarious calibration of Earth-observing sensors. RadCaTS was upgraded to use RSG custom 8-band ground viewing radiometers (GVRs) beginning in 2011 and currently four GVRs are deployed providing an average reflectance for the test site. This measurement of ground reflectance is the most critical component of vicarious calibration using the reflectance-based method. In order to ensure the quality of these measurements, RSG has been exploring more efficient and accurate methods of on-site calibration evaluation. This work describes the design of, and initial results from, a small portable transfer radiometer for the purpose of GVR calibration validation on site. Prior to deployment, RSG uses high accuracy laboratory calibration methods in order to provide radiance calibrations with low uncertainties for each GVR. After deployment, a solar radiation based calibration has typically been used. The method is highly dependent on a clear, stable atmosphere, requires at least two people to perform, is time consuming in post processing, and is dependent on several large pieces of equipment. In order to provide more regular and more accurate calibration monitoring, the small portable transfer radiometer is designed for quick, one-person operation and on-site field calibration comparison results. The radiometer is also suited for laboratory calibration use and thus could be used as a transfer radiometer calibration standard for ground viewing radiometers of a RadCalNet site.

  18. Calibration of Discrete Element Heat Transfer Parameters by Central Composite Design

    Science.gov (United States)

    Deng, Zongquan; Cui, Jinsheng; Hou, Xuyan; Jiang, Shengyuan

    2017-03-01

    The efficiency and precision of parameter calibration in discrete element method (DEM) are not satisfactory, and parameter calibration for granular heat transfer is rarely involved. Accordingly, parameter calibration for granular heat transfer with the DEM is studied. The heat transfer in granular assemblies is simulated with DEM, and the effective thermal conductivity (ETC) of these granular assemblies is measured with the transient method in simulations. The measurement testbed is designed to test the ETC of the granular assemblies under normal pressure and a vacuum based on the steady method. Central composite design (CCD) is used to simulate the impact of the DEM parameters on the ETC of granular assemblies, and the heat transfer parameters are calibrated and compared with experimental data. The results show that, within the scope of the considered parameters, the ETC of the granular assemblies increases with an increasing particle thermal conductivity and decreases with an increasing particle shear modulus and particle diameter. The particle thermal conductivity has the greatest impact on the ETC of granular assemblies followed by the particle shear modulus and then the particle diameter. The calibration results show good agreement with the experimental results. The error is less than 4%, which is within a reasonable range for the scope of the CCD parameters. The proposed research provides high efficiency and high accuracy parameter calibration for granular heat transfer in DEM.

  19. Calibrating GPS With TWSTFT For Accurate Time Transfer

    Science.gov (United States)

    2008-12-01

    and O. Koudelka, 2008, “Time transfer with nanosecond accuracy for the realization of International Atomic Time,” Metrologia , 45, 185- 198. [4] Z...468-475. [7] Z. Jiang, 2008, “Towards a TWSTFT Network Time Transfer,” Metrologia , 45, S6-S11.

  20. A conjugate heat transfer procedure for gas turbine blades.

    Science.gov (United States)

    Croce, G

    2001-05-01

    A conjugate heat transfer procedure, allowing for the use of different solvers on the solid and fluid domain(s), is presented. Information exchange between solid and fluid solution is limited to boundary condition values, and this exchange is carried out at any pseudo-time step. Global convergence rate of the procedure is, thus, of the same order of magnitude of stand-alone computations.

  1. Research on Calibration Transfer across Infrared Spectrometric Instruments for Fast Evaluation of Crude Oils

    Institute of Scientific and Technical Information of China (English)

    Li Jingyan; Chu Xiaoli; Tian Songbai

    2015-01-01

    The spectral analysis method is suitable for the process control and the process analysis such as the fast evalua-tion of crude oils. In this study, model transfer between the same type of spectrometer and different type of spectrometers was discussed respectively, and the results have shown that the transfer of calibration model can satisfy the demand of rapid analysis.

  2. Enhancing the quality of hydrologic model calibrations and their transfer to operational flood forecasters

    Science.gov (United States)

    Aggett, Graeme; Spies, Ryan; Szfranski, Bill; Hahn, Claudia; Weil, Page

    2016-04-01

    An adequate forecasting model may not perform well if it is inadequately calibrated. Model calibration is often constrained by the lack of adequate calibration data, especially for small river basins with high spatial rainfall variability. Rainfall/snow station networks may not be dense enough to accurately estimate the catchment rainfall/SWE. High discharges during flood events are subject to significant error due to flow gauging difficulty. Dynamic changes in catchment conditions (e.g., urbanization; losses in karstic systems) invariably introduce non-homogeneity in the water level and flow data. This presentation will highlight some of the challenges in reliable calibration of National Weather Service (i.e. US) operational flood forecast models, emphasizing the various challenges in different physiographic/climatic domains. It will also highlight the benefit of using various data visualization techniques to transfer information about model calibration to operational forecasters so they may understand the influence of the calibration on model performance under various conditions.

  3. A calibration procedure to improve global rice yield simulations with EPIC

    Energy Technology Data Exchange (ETDEWEB)

    Xiong, Wei; Balkovic, Juraj; van der Velde, M.; Zhang, Xuesong; Izaurralde, Roberto C.; Skalsky, Rastislav; Lin, Erda; Mueller, Nathan; Obersteiner, Michael

    2014-02-01

    Crop models are increasingly used to assess impacts of climate change/variability and management practices on productivity and environmental performance of alternative cropping systems. Calibration is an important procedure to improve reliability of model simulations, especially for large area applications. However, global-scale crop model calibration has rarely been exercised due to limited data availability and expensive computing cost. Here we present a simple approach to calibrate Environmental Policy Integrated Climate (EPIC) model for a global implementation of rice. We identify four parameters (potential heat unit – PHU, planting density – PD, harvest index – HI, and biomass energy ratio – BER) and calibrate them regionally to capture the spatial pattern of reported rice yield in 2000. Model performance is assessed by comparing simulated outputs with independent FAO national data. The comparison demonstrates that the global calibration scheme performs satisfactorily in reproducing the spatial pattern of rice yield, particularly in main rice production areas. Spatial agreement increases substantially when more parameters are selected and calibrated, but with varying efficiencies. Among the parameters, PHU and HI exhibit the highest efficiencies in increasing the spatial agreement. Simulations with different calibration strategies generate a pronounced discrepancy of 5–35% in mean yields across latitude bands, and a small to moderate difference in estimated yield variability and yield changing trend for the period of 1981–2000. Present calibration has little effects in improving simulated yield variability and trends at both regional and global levels, suggesting further works are needed to reproduce temporal variability of reported yields. This study highlights the importance of crop models’ calibration, and presents the possibility of a transparent and consistent up scaling approach for global crop simulations given current availability of

  4. Functional calibration procedure for 3D knee joint angle description using inertial sensors.

    Science.gov (United States)

    Favre, J; Aissaoui, R; Jolles, B M; de Guise, J A; Aminian, K

    2009-10-16

    Measurement of three-dimensional (3D) knee joint angle outside a laboratory is of benefit in clinical examination and therapeutic treatment comparison. Although several motion capture devices exist, there is a need for an ambulatory system that could be used in routine practice. Up-to-date, inertial measurement units (IMUs) have proven to be suitable for unconstrained measurement of knee joint differential orientation. Nevertheless, this differential orientation should be converted into three reliable and clinically interpretable angles. Thus, the aim of this study was to propose a new calibration procedure adapted for the joint coordinate system (JCS), which required only IMUs data. The repeatability of the calibration procedure, as well as the errors in the measurement of 3D knee angle during gait in comparison to a reference system were assessed on eight healthy subjects. The new procedure relying on active and passive movements reported a high repeatability of the mean values (offset0.9). In comparison to the reference system, this functional procedure showed high precision (SD0.75) and moderate accuracy (between 4.0 degrees and 8.1 degrees) for the three knee angle. The combination of the inertial-based system with the functional calibration procedure proposed here resulted in a promising tool for the measurement of 3D knee joint angle. Moreover, this method could be adapted to measure other complex joint, such as ankle or elbow.

  5. L5 TM radiometric recalibration procedure using the internal calibration trends from the NLAPS trending database

    Science.gov (United States)

    Chander, G.; Haque, Md. O.; Micijevic, E.; Barsi, J.A.

    2008-01-01

    From the Landsat program's inception in 1972 to the present, the earth science user community has benefited from a historical record of remotely sensed data. The multispectral data from the Landsat 5 (L5) Thematic Mapper (TM) sensor provide the backbone for this extensive archive. Historically, the radiometric calibration procedure for this imagery used the instrument's response to the Internal Calibrator (IC) on a scene-by-scene basis to determine the gain and offset for each detector. The IC system degraded with time causing radiometric calibration errors up to 20 percent. In May 2003 the National Landsat Archive Production System (NLAPS) was updated to use a gain model rather than the scene acquisition specific IC gains to calibrate TM data processed in the United States. Further modification of the gain model was performed in 2007. L5 TM data that were processed using IC prior to the calibration update do not benefit from the recent calibration revisions. A procedure has been developed to give users the ability to recalibrate their existing Level-1 products. The best recalibration results are obtained if the work order report that was originally included in the standard data product delivery is available. However, many users may not have the original work order report. In such cases, the IC gain look-up table that was generated using the radiometric gain trends recorded in the NLAPS database can be used for recalibration. This paper discusses the procedure to recalibrate L5 TM data when the work order report originally used in processing is not available. A companion paper discusses the generation of the NLAPS IC gain and bias look-up tables required to perform the recalibration.

  6. Calibration transfer based on maximum margin criterion for qualitative analysis using Fourier transform infrared spectroscopy.

    Science.gov (United States)

    Hu, Yong; Peng, Silong; Bi, Yiming; Tang, Liang

    2012-12-21

    A traditional multivariate calibration transfer method such as piecewise direct standardization (PDS) is usually applied to quantitative analysis. To make the method apply to qualitative analysis of Fourier Transform Infrared spectroscopy (FTIR), we propose an improved calibration transfer method based on the maximum margin criterion (CTMMC). The new method not only considers the spectral changes under different conditions, but also takes into account the geometric characteristics of spectra from different classes, so the transformed spectra from different classes will be separated as far as possible, and this will improve the performance of the follow-up qualitative analysis. A comparative study is provided between the proposed method CTMMC and other traditional calibration transfer methods on two data sets. Experimental results show that the proposed method can achieve better performance than previous methods.

  7. Transferability of calibrated microsimulation model parameters for safety assessment using simulated conflicts.

    Science.gov (United States)

    Essa, Mohamed; Sayed, Tarek

    2015-11-01

    Several studies have investigated the relationship between field-measured conflicts and the conflicts obtained from micro-simulation models using the Surrogate Safety Assessment Model (SSAM). Results from recent studies have shown that while reasonable correlation between simulated and real traffic conflicts can be obtained especially after proper calibration, more work is still needed to confirm that simulated conflicts provide safety measures beyond what can be expected from exposure. As well, the results have emphasized that using micro-simulation model to evaluate safety without proper model calibration should be avoided. The calibration process adjusts relevant simulation parameters to maximize the correlation between field-measured and simulated conflicts. The main objective of this study is to investigate the transferability of calibrated parameters of the traffic simulation model (VISSIM) for safety analysis between different sites. The main purpose is to examine whether the calibrated parameters, when applied to other sites, give reasonable results in terms of the correlation between the field-measured and the simulated conflicts. Eighty-three hours of video data from two signalized intersections in Surrey, BC were used in this study. Automated video-based computer vision techniques were used to extract vehicle trajectories and identify field-measured rear-end conflicts. Calibrated VISSIM parameters obtained from the first intersection which maximized the correlation between simulated and field-observed conflicts were used to estimate traffic conflicts at the second intersection and to compare the results to parameters optimized specifically for the second intersection. The results show that the VISSIM parameters are generally transferable between the two locations as the transferred parameters provided better correlation between simulated and field-measured conflicts than using the default VISSIM parameters. Of the six VISSIM parameters identified as

  8. A C-arm calibration method with application to fluoroscopic image-guided procedures

    Science.gov (United States)

    Rai, Lav; Gibbs, Jason D.; Wibowo, Henky

    2012-02-01

    C-arm fluoroscopy units provide continuously updating X-ray video images during surgical procedure. The modality is widely adopted for its low cost, real-time imaging capabilities, and its ability to display radio-opaque tools in the anatomy. It is, however, important to correct for fluoroscopic image distortion and estimate camera parameters, such as focal length and camera center, for registration with 3D CT scans in fluoroscopic imageguided procedures. This paper describes a method for C-arm calibration and evaluates its accuracy in multiple C-arm units and in different viewing orientations. The proposed calibration method employs a commerciallyavailable unit to track the C-arm and a calibration plate. The method estimates both the internal calibration parameters and the transformation between the coordinate systems of tracker and C-arm. The method was successfully tested on two C-arm units (GE OEC 9800 and GE OEC 9800 Plus) of different image intensifier sizes and verified with a rigid airway phantom model. The mean distortion-model error was found to be 0.14 mm and 0.17 mm for the respective C-arms. The mean overall system reprojection error (which measures the accuracy of predicting an image using tracker coordinates) was found to be 0.63 mm for the GE OEC 9800.

  9. Reassessing the MADE direct-push hydraulic conductivity data using a revised calibration procedure

    Science.gov (United States)

    Bohling, Geoffrey C.; Liu, Gaisheng; Dietrich, Peter; Butler, James J.

    2016-11-01

    In earlier work, we presented a geostatistical assessment of high-resolution hydraulic conductivity (K) profiles obtained at the MADE site using direct-push (DP) methods. The profiles are derived from direct-push injection logger (DPIL) measurements that provide a relative indicator of vertical variations in K with a sample spacing of 1.5 cm. The DPIL profiles are converted to K profiles by calibrating to the results of direct-push permeameter (DPP) tests performed at selected depths in some of the profiles. Our original calibration used a linear transform that failed to adequately account for an upper limit on DPIL responses in high-K zones and noise in the DPIL data. Here we present a revised calibration procedure that accounts for the upper limit and noise, leading to DPIL K values that display a somewhat different univariate distribution and a lower lnK variance (5.9 ± 1.5) than the original calibration values (6.9 ± 1.8), although each variance estimate falls within the other's 95% confidence interval. Despite the change in the univariate distribution, the autocorrelation structure and large-scale patterns exhibited by the revised DPIL K values still agree well with those exhibited by the flowmeter data from the site. We provide the DPIL and DPP data, along with our calibrated DPIL K values, in the Supporting Information.

  10. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: expanding the application range

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy V.; Anderson, Erik H.; Barber, Samuel K.; Bouet, Nathalie; Cambie, Rossana; Conley, Raymond; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitriy L.

    2011-03-14

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays [Proc. SPIE 7077-7 (2007), Opt. Eng. 47, 073602 (2008)] has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer [Nucl. Instr. and Meth. A616, 172 (2010)]. Here we report on a further expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  11. Development and calibration of heat transfer coefficient sensor for rotational state

    Institute of Scientific and Technical Information of China (English)

    QUAN Yong-kai; XU Guo-qiang; LUO Xiang; ZHANG Da

    2011-01-01

    A novel heat transfer coefficient sensor is introduced and the design,manufacture,and calibration are described.The intended application of this instrument was on a high rotational speed test disc.In the experiments,the heat transfer coefficient sensor was calibrated under static state and rotational state respectively.The calibration under the static state was accomplished in a pipe: the inside diameter(ID) was 0.048m and the total length was 4m,the distance between the sensor and the inlet of the pipe was 3.5m;the standard value was measured using a self-made calibrator.The calibration under the rotational state was accomplished using a rotating disc: the diameter and thickness of the stainless disc were 800mm and 25mm,respectively;the sensor was installed at the location of r=250mm;the rotating disc driven by a 30kW direct current motor can supply the maximum rotational speed of 3000r/min.The standard value under rotational state was provided by an approximate empirical formula.The results show that the designed sensor can measure heat transfer coefficient directly under rotational state and static state with good accuracy and stability;the correlation factor of K are constant under static state and rotational state.

  12. Calibration of a Numerical Model for Heat Transfer and Fluid Flow in an Extruder

    DEFF Research Database (Denmark)

    Hofstätter, Thomas; Pedersen, David Bue; Nielsen, Jakob Skov

    2016-01-01

    This paper discusses experiments performed in order to validate simulations on a fused deposition modelling (FDM) extruder. The nozzle has been simulated in terms of heat transfer and fluid flow. In order to calibrate and validate these simulations, experiments were performed giving a significant...

  13. Calibration procedures for imaging spectrometers: improving data quality from satellite missions to UAV campaigns

    Science.gov (United States)

    Brachmann, Johannes F. S.; Baumgartner, Andreas; Lenhard, Karim

    2016-10-01

    The Calibration Home Base (CHB) at the Remote Sensing Technology Institute of the German Aerospace Center (DLR-IMF) is an optical laboratory designed for the calibration of imaging spectrometers for the VNIR/SWIR wavelength range. Radiometric, spectral and geometric characterization is realized in the CHB in a precise and highly automated fashion. This allows performing a wide range of time consuming measurements in an efficient way. The implementation of ISO 9001 standards ensures a traceable quality of results. DLR-IMF will support the calibration and characterization campaign of the future German spaceborne hyperspectral imager EnMAP. In the context of this activity, a procedure for the correction of imaging artifacts, such as due to stray light, is currently being developed by DLR-IMF. Goal is the correction of in-band stray light as well as ghost images down to a level of a few digital numbers in the whole wavelength range 420-2450 nm. DLR-IMF owns a Norsk Elektro Optikks HySpex airborne imaging spectrometer system that has been thoroughly characterized. This system will be used to test stray light calibration procedures for EnMAP. Hyperspectral snapshot sensors offer the possibility to simultaneously acquire hyperspectral data in two dimensions. Recently, these rather new spectrometers have arisen much interest in the remote sensing community. Different designs are currently used for local area observation such as by use of small unmanned aerial vehicles (sUAV). In this context the CHB's measurement capabilities are currently extended such that a standard measurement procedure for these new sensors will be implemented.

  14. Calibration procedures for improved accuracy of wind turbine blade load measurement

    Energy Technology Data Exchange (ETDEWEB)

    Dahlberg, J.Aa. [Aeronautical Research Inst. of Sweden, Bromma (Sweden); Johansson, Hjalmar [Teknikgruppen AB, Sollentuna (Sweden)

    1996-12-01

    External loads acting on wind turbine blades are mainly transferred via the hub to the rest of the structure. It is therefore a normal approach to measure the loads acting on the turbine by load measurements in the blade roots. The load measurement is often accomplished by measurements of strain on the surface of the blade or the hub. The strain signals are converted to loads by applying calibration factors to the measurements. This paper deals with difficulties associated with load measurements on two different wind turbines; one with strain gauges applied to a steel hub where a linear stress-load relationship is expected and the other with strain gauges applied to the GFRP blade close to the bearings where strong non-linearity`s and temperature effects are expected. This paper suggests calibration methods to overcome these problems. 2 refs, 11 figs

  15. 40 CFR Appendix C to Part 50 - Measurement Principle and Calibration Procedure for the Measurement of Carbon Monoxide in the...

    Science.gov (United States)

    2010-07-01

    ... susceptibility to interference, malfunction, or damage due to particles. Calibration Procedure 1. Principle... connection. The system must have a vent designed to insure atmospheric pressure at the manifold and...

  16. Archimedean Spiral Antenna Calibration Procedures to Increase the Downrange Resolution of a SFCW Radar

    Directory of Open Access Journals (Sweden)

    Ioan Nicolaescu

    2008-01-01

    Full Text Available This paper deals with the calibration procedures of an Archimedean spiral antenna used for a stepped frequency continuous wave radar (SFCW, which works from 400 MHz to 4845 MHz. Two procedures are investigated, one based on an error-term flow graph for the frequency signal and the second based on a reference metallic plate located at a certain distance from the ground in order to identify the phase dispersion given by the antenna. In the second case, the received signal is passed in time domain by applying an ifft, the multiple reflections are removed and the phase variation due to the time propagation is subtracted. After phase correction, the time domain response as well as the side lobes level is decreased. The antenna system made up of two Archimedean spirals is employed by SFCW radar that operates with a frequency step of 35 MHz.

  17. Competence of Testing and Calibrating Laboratories, Procedure to Facilitate the Process.

    Directory of Open Access Journals (Sweden)

    Ivonne Jacqueline Valdés Peña

    2013-03-01

    Full Text Available As an investigation in the field of Laboratories competence, the proposal of a an evaluation procedure and another one to facilitate the competence process of calibrating and testing labs, permitted to identify the main causes that influence on this process using basic quality tools such as: brain storming, cause-effect diagram, Pareto diagram and surveys. The result of one stratific pattern for organism compiled the 94 % of the laboratories surveyed it is showed the need of focussing the efforts on the categories: Personnel, managers´ training about the standards for competence, selection of responsible quality personnel, training of lab technicians; Methods, home auditing Program and documented procedures that assure the quality results. Based on these causes are stated the reasons that create the basis to Laboratories competence in Sancti Spíritus, that until this moment is missing a methodological instrument to face this process.

  18. An alternative method for calibration of narrow band radiometer using a radiative transfer model

    Energy Technology Data Exchange (ETDEWEB)

    Salvador, J; Wolfram, E; D' Elia, R [Centro de Investigaciones en Laseres y Aplicaciones, CEILAP (CITEFA-CONICET), Juan B. de La Salle 4397 (B1603ALO), Villa Martelli, Buenos Aires (Argentina); Zamorano, F; Casiccia, C [Laboratorio de Ozono y Radiacion UV, Universidad de Magallanes, Punta Arenas (Chile) (Chile); Rosales, A [Universidad Nacional de la Patagonia San Juan Bosco, UNPSJB, Facultad de Ingenieria, Trelew (Argentina) (Argentina); Quel, E, E-mail: jsalvador@citefa.gov.ar [Universidad Nacional de la Patagonia Austral, Unidad Academica Rio Gallegos Avda. Lisandro de la Torre 1070 ciudad de Rio Gallegos-Sta Cruz (Argentina) (Argentina)

    2011-01-01

    The continual monitoring of solar UV radiation is one of the major objectives proposed by many atmosphere research groups. The purpose of this task is to determine the status and degree of progress over time of the anthropogenic composition perturbation of the atmosphere. Such changes affect the intensity of the UV solar radiation transmitted through the atmosphere that then interacts with living organisms and all materials, causing serious consequences in terms of human health and durability of materials that interact with this radiation. One of the many challenges that need to be faced to perform these measurements correctly is the maintenance of periodic calibrations of these instruments. Otherwise, damage caused by the UV radiation received will render any one calibration useless after the passage of some time. This requirement makes the usage of these instruments unattractive, and the lack of frequent calibration may lead to the loss of large amounts of acquired data. Motivated by this need to maintain calibration or, at least, know the degree of stability of instrumental behavior, we have developed a calibration methodology that uses the potential of radiative transfer models to model solar radiation with 5% accuracy or better relative to actual conditions. Voltage values in each radiometer channel involved in the calibration process are carefully selected from clear sky data. Thus, tables are constructed with voltage values corresponding to various atmospheric conditions for a given solar zenith angle. Then we model with a radiative transfer model using the same conditions as for the measurements to assemble sets of values for each zenith angle. The ratio of each group (measured and modeled) allows us to calculate the calibration coefficient value as a function of zenith angle as well as the cosine response presented by the radiometer. The calibration results obtained by this method were compared with those obtained with a Brewer MKIII SN 80 located in the

  19. Calibration of a Numerical Model for Heat Transfer and Fluid Flow in an Extruder

    DEFF Research Database (Denmark)

    Hofstätter, Thomas; Pedersen, David Bue; Nielsen, Jakob Skov

    2016-01-01

    This paper discusses experiments performed in order to validate simulations on a fused deposition modelling (FDM) extruder. The nozzle has been simulated in terms of heat transfer and fluid flow. In order to calibrate and validate these simulations, experiments were performed giving a significant...... dynamical parameters. This research sets the foundation for further research within melted extrusion based additive manufacturing. The heating process of the extruder will be described and a note on the material feeding will be given....

  20. Stability of modulation transfer function calibration of surface profilometers using binary pseudo-random gratings and arrays with nonideal groove shapes

    Energy Technology Data Exchange (ETDEWEB)

    Barber, Samuel K.; Anderson, Erik H.; Cambie, Rossana; Marchesini, Stefano; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitry L.; Yashchuk, Valeriy V.

    2010-03-31

    The major problem of measurement of a power spectral density (PSD) distribution of surface heights with surface profilometers arises due to the unknown Modulation Transfer Function (MTF) of the instruments, which tends to distort the PSD at higher spatial frequencies. The special mathematical properties of binary pseudo-random patterns make them an ideal basis for developing MTF calibration test surfaces. Two-dimensional binary pseudo-random arrays (BPRAs) have been fabricated and used for the MTF calibration of the MicroMap{trademark}-570 interferometric microscope with all available objectives. An investigation into the effects of fabrication imperfections on the quality of the MTF calibration and a procedure for accounting for such imperfections are presented.

  1. Development of a procedure of calibration of meters of product Kerma-Area; Desarrollo de un procedimiento de calibracion de medidores de producto Kerma-Area

    Energy Technology Data Exchange (ETDEWEB)

    Ginjaume, M.; Roig, M.; Amores, M.; Ortega, X.

    2013-07-01

    The aim of this paper is to present the calibration procedure developed, uncertainties associated and scope of accredited degree. Also the objectives of EURAMET 1177 and calibrations are described conducted by our laboratory in this intercomparison. (Author)

  2. Time transfer by laser link (T2L2): characterization and calibration of the flight instrument

    Science.gov (United States)

    Samain, E.; Vrancken, P.; Guillemot, P.; Fridelance, P.; Exertier, P.

    2014-10-01

    The T2L2 project (time transfer by laser link) allows for the synchronization of remote ultra-stable clocks over intercontinental distances (Fridelance et al 1997 Exp. Astron. 7, Samain and Fridelance 1998 Metrologia 35 151-9). The principle is derived from satellite laser ranging technology with dedicated space equipment designed to record arrival times of laser pulses at the satellite. The space segment has been launched in June 2008 as a passenger experiment on the ocean altimetry satellite Jason 2. T2L2 had been specified to yield a time stability of better than 1 ps over 1000 s integration time and an accuracy of better than 100 ps. This level of performance requires a rigorous data processing which can be performed only with a comprehensive calibration model of the whole instrumentation. For this purpose, several experimental measurements have been performed before and during the integration phase of the T2L2 space instrument. This instrument model is one of the cornerstones of the data reduction process which is carried out to translate the raw information to a usable picosecond time transfer. After providing a global synopsis of the T2L2 space instrument, the paper gives a description of the experimental setup for the instrument characterization. It then details the different contributions within the calibration model and concludes with an applied example of a space to ground time transfer.

  3. Design, calibration and error analysis of instrumentation for heat transfer measurements in internal combustion engines

    Science.gov (United States)

    Ferguson, C. R.; Tree, D. R.; Dewitt, D. P.; Wahiduzzaman, S. A. H.

    1987-01-01

    The paper reports the methodology and uncertainty analyses of instrumentation for heat transfer measurements in internal combustion engines. Results are presented for determining the local wall heat flux in an internal combustion engine (using a surface thermocouple-type heat flux gage) and the apparent flame-temperature and soot volume fraction path length product in a diesel engine (using two-color pyrometry). It is shown that a surface thermocouple heat transfer gage suitably constructed and calibrated will have an accuracy of 5 to 10 percent. It is also shown that, when applying two-color pyrometry to measure the apparent flame temperature and soot volume fraction-path length, it is important to choose at least one of the two wavelengths to lie in the range of 1.3 to 2.3 micrometers. Carefully calibrated two-color pyrometer can ensure that random errors in the apparent flame temperature and in the soot volume fraction path length will remain small (within about 1 percent and 10-percent, respectively).

  4. Calibration chain design based on integrating sphere transfer radiometer for SI-traceable on-orbit spectral radiometric calibration and its uncertainty analysis

    Science.gov (United States)

    Zhao, Wei-Ning; Fang, Wei; Sun, Li-Wei; Cui, Li-Hong; Wang, Yu-Peng

    2016-09-01

    In order to satisfy the requirement of SI-traceable on-orbit absolute radiation calibration transfer with high accuracy for satellite remote sensors, a transfer chain consisting of a fiber coupling monochromator (FBM) and an integrating sphere transfer radiometer (ISTR) was designed in this paper. Depending on the Sun, this chain based on detectors provides precise spectral radiometric calibration and measurement to spectrometers in the reflective solar band (RSB) covering 300-2500 nm with a spectral bandwidth of 0.5-6 nm. It shortens the traditional chain based on lamp source and reduces the calibration uncertainty from 5% to 0.5% by using the cryogenic radiometer in space as a radiometric benchmark and trap detectors as secondary standard. This paper also gives a detailed uncertainty budget with reasonable distribution of each impact factor, including the weak spectral signal measurement with uncertainty of 0.28%. According to the peculiar design and comprehensive uncertainty analysis, it illustrates that the spectral radiance measurement uncertainty of the ISTR system can reach to 0.48%. The result satisfies the requirements of SI-traceable on-orbit calibration and has wider significance for expanding the application of the remote sensing data with high-quality. Project supported by the National Natural Science Foundation of China (Grant No. 41474161) and the National High-Technology Program of China (Grant No. 2015AA123703).

  5. Validation tests of open-source procedures for digital camera calibration and 3D image-based modelling

    OpenAIRE

    I. Toschi; Rivola, R.; Bertacchini, E; Castagnetti, C.; M. Dubbini; Capra, A.

    2013-01-01

    Among the many open-source software solutions recently developed for the extraction of point clouds from a set of un-oriented images, the photogrammetric tools Apero and MicMac (IGN, Institut Géographique National) aim to distinguish themselves by focusing on the accuracy and the metric content of the final result. This paper firstly aims at assessing the accuracy of the simplified and automated calibration procedure offered by the IGN tools. Results obtained with this procedure were...

  6. Calibration and intercomparison of acetic acid measurements using proton transfer reaction mass spectrometry (PTR-MS

    Directory of Open Access Journals (Sweden)

    K. B. Haase

    2012-07-01

    Full Text Available Acetic acid is one of the most abundant organic acids in the ambient atmosphere, with maximum mixing ratios reaching into the tens of parts per billion by volume (ppbv range. The identities and associated magnitudes of the major sources and sinks for acetic acid are poorly characterized, due in part to the limitation in available measurement techniques. This paper demonstrates that Proton Transfer Reaction Mass Spectrometry (PTR-MS can reliably quantify acetic acid vapor in ambient air. Three different PTR-MS configurations were calibrated at low ppbv mixing ratios using permeation tubes, which yielded calibration factors between 7.0 and 10.9 normalized counts per second per ppbv (ncps ppbv−1 at a drift tube field strength of 132 townsend (Td. Detection limits ranged from 0.06 to 0.32 ppbv with dwell times of 5 s. These calibration factors showed negligible humidity dependence. Using the experimentally determined calibration factors, PTR-MS measurements of acetic acid during the International Consortium for Atmospheric Research on Transport and Transformation (ICARTT campaign were validated against results obtained using Mist Chambers coupled with Ion Chromatography (MC/IC. An orthogonal least squares linear regression of paired data yielded a slope of 1.14 ± 0.06 (2σ, an intercept of 0.049 ± 20 (2σ ppbv, and an R2 of 0.78. The median mixing ratio of acetic acid on Appledore Island, ME during the ICARTT campaign was 0.530 ± 0.025 ppbv with a minimum of 0.075 ± 0.004 ppbv, and a maximum of 3.555 ± 0.171 ppbv.

  7. Calibration and intercomparison of acetic acid measurements using proton transfer reaction mass spectrometry (PTR-MS)

    Science.gov (United States)

    Haase, K.B.; Keene, W.C.; Pszenny, A.A.P.; Mayne, H.R.; Talbot, R.W.; Sive, B.C.

    2012-01-01

    Acetic acid is one of the most abundant organic acids in the ambient atmosphere, with maximum mixing ratios reaching into the tens of parts per billion by volume (ppbv) range. The identities and associated magnitudes of the major sources and sinks for acetic acid are poorly characterized, due in part to the limitation in available measurement techniques. This paper demonstrates that Proton Transfer Reaction Mass Spectrometry (PTR-MS) can reliably quantify acetic acid vapor in ambient air. Three different PTR-MS configurations were calibrated at low ppbv mixing ratios using permeation tubes, which yielded calibration factors between 7.0 and 10.9 normalized counts per second per ppbv (ncps ppbv−1) at a drift tube field strength of 132 townsend (Td). Detection limits ranged from 0.06 to 0.32 ppbv with dwell times of 5 s. These calibration factors showed negligible humidity dependence. Using the experimentally determined calibration factors, PTR-MS measurements of acetic acid during the International Consortium for Atmospheric Research on Transport and Transformation (ICARTT) campaign were validated against results obtained using Mist Chambers coupled with Ion Chromatography (MC/IC). An orthogonal least squares linear regression of paired data yielded a slope of 1.14 ± 0.06 (2σ), an intercept of 0.049 ± 20 (2σ) ppbv, and an R2 of 0.78. The median mixing ratio of acetic acid on Appledore Island, ME during the ICARTT campaign was 0.530 ± 0.025 ppbv with a minimum of 0.075 ± 0.004 ppbv, and a maximum of 3.555 ± 0.171 ppbv.

  8. A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

    Science.gov (United States)

    Fidanzio, Andrea; Greco, Francesca; Gargiulo, Laura; Cilla, Savino; Sabatino, Domenico; Cappiello, Massimo; Di Felice, Cinzia; Di Castro, Elisabetta; Azario, Luigi; Russo, Mariateresa; Pompei, Luciano; D'Onofrio, Guido; Piermattei, Angelo

    2011-03-01

    A practical and accurate generalized in vivo dosimetry procedure has been implemented for Siemens linacs supplying 6, 10, and 15 MV photon beams, equipped with aSi electronic portal imaging devices (EPIDs). The in vivo dosimetry method makes use of correlation ratios between EPID transit signal, s (t) (0) (TPR,w,L), and phantom mid-plane dose, D (0)(TPR,w,L), as functions of phantom thickness, w, square field dimensions, L, and tissue-phantom ratio TPR(20,10). The s (t) (0) (TPR,w,L) and D (0)(TPR,w,L) values were defined to be independent of the EPID sensitivity and monitor unit calibration, while their dependence on TPR(20,10) was investigated to determine a set of generalized correlation ratios to be used for beams with TPR(20,10) falling in the examined range. This way, other radiotherapy centers can use the method with no need to locally perform the whole set of measurements in solid water phantoms, required to implement it. Tolerance levels for 3D conformal treatments, ranging between ±5 and ±6% according to tumor type and location, were estimated for comparison purposes between reconstructed isocenter dose, D (iso), and treatment planning system (TPS) computed dose D (iso,TPS). Finally a dedicated software, interfaceable with record and verify (R&V) systems used in the centers, was developed to obtain in vivo dosimetry results in less than 2 min after beam delivery.

  9. Clinic based transfer of the N(D,W)(60Co) calibration coefficient using a linear accelerator.

    Science.gov (United States)

    Abdel-Rahman, W; Evans, M D C; Serré, L; McCaffrey, J P; Podgorsak, E B; Seuntjens, J P

    2009-03-01

    Ionization chambers used for reference dosimetry require a local secondary standard ionization chamber with a 60Co absorbed dose to water calibration coefficient N(D,W)(60Co) traceable to a national primary standards dosimetry laboratory or an accredited secondary dosimetry calibration laboratory. Clinic based (in-house) transfer of this coefficient to tertiary reference ionization chambers has traditionally been accomplished with chamber cross calibration in water using a 60Co beam; however, access to 60Co teletherapy machines has become increasingly limited for clinic based physicists. In this work, the accuracy of alternative methods of transferring the N(D,W)(60Co) calibration coefficient using 6 and 18 MV photon beams from a linear accelerator in lieu of 60Co has been investigated for five different setups and four commonly used chamber types.

  10. Novel Real-time Calibration and Alignment Procedure for LHCb Run II

    CERN Multimedia

    Prouve, Claire

    2016-01-01

    In order to achieve optimal detector performance the LHCb experiment has introduced a novel real-time detector alignment and calibration strategy for Run II of the LHC. For the alignment tasks, data is collected and processed at the beginning of each fill while the calibrations are performed for each run. This real time alignment and calibration allows the same constants being used in both the online and offline reconstruction, thus improving the correlation between triggered and offline selected events. Additionally the newly computed alignment and calibration constants can be instantly used in the trigger, making it more efficient. The online alignment and calibration of the RICH detectors also enable the use of hadronic particle identification in the trigger. The computing time constraints are met through the use of a new dedicated framework using the multi-core farm infrastructure for the LHCb trigger. An overview of all alignment and calibration tasks is presented and their performance is shown.

  11. A direct comparison between two independently calibrated time transfer techniques: T2L2 and GPS Common-Views

    Science.gov (United States)

    Rovera, G. D.; Abgrall, M.; Courde, C.; Exertier, P.; Fridelance, P.; Guillemot, Ph; Laas-Bourez, M.; Martin, N.; Samain, E.; Sherwood, R.; Torre, J.-M.; Uhrich, P.

    2016-06-01

    We present a direct comparison between two satellite time transfer techniques on independently calibrated links: Time Transfer by Laser Link (T2L2) and Common-Views (CV) of satellites from the Global Positioning System (GPS) constellation. The GPS CV and T2L2 links between three European laboratories where independently calibrated against the same reference point of the local timescales. For all the links the mean values of the differences between GPS CV and T2L2 are equal or below 240 ps, with standard deviations below 500 ps, mostly due to GPS CV. Almost all deviations from 0 ns are within the combined uncertainty estimates. Despite the weak number of common points obtained, due to the fact that T2L2 is weather dependent, these results are providing an unprecedented sub-ns consistency between two independently calibrated microwave and optical satellite time transfer techniques.

  12. Evaluation of several calibration procedures for a portable soil moisture sensor

    Science.gov (United States)

    Rowlandson, Tracy L.; Berg, Aaron A.; Bullock, Paul R.; Ojo, E. RoTimi; McNairn, Heather; Wiseman, Grant; Cosh, Michael H.

    2013-08-01

    The calibration and validation of remotely sensed soil moisture products relies upon an accurate source of ground truth data. The primary method of providing this ground truth is to conduct intensive field campaigns with manual surface soil moisture sampling measurements, which utilize gravimetric sampling, soil moisture probes, or both, to estimate the volumetric soil water content. Soil moisture probes eliminate the need for labor-intensive gravimetric sampling. To ensure the accuracy of these probes, several studies have determined these probes need various degrees of localized calibration. This study examines six possible calibration techniques using data collected during a field campaign conducted in 2012, with soil moisture samples being collected over 55 fields in southern Manitoba, as part of the Soil Moisture Active Passive Validation Experiment 2012 (SMAPVEX12). The use of a general equation, applied to all collected data, resulted in the largest error regardless of whether a linear or third order polynomial relationship was established for the calibration of the soil moisture probes. Calibration equations based on soil texture or vegetation land cover reduced the error; however, the individual calibration equations established for each field in the study had the lowest error of all the calibration techniques. Although average bias was low for all of the calibration techniques, the use of the general equation to calibrate individual fields resulted in high biases for some fields.

  13. Evaluation of debris flow susceptibility by means of a transferability procedure: a study case in Messina area

    Science.gov (United States)

    Cama, Mariaelena; Luigi, Lombardo; Conoscenti, Christian; Rotigliano, Edoardo

    2016-04-01

    Debris flows can be described as rapid mass movements, gravity induced able to transport large quantities of material downslope. This type of landslides is strongly controlled by the topography and usually occur in region characterized by steep slopes and at least seasonal heavy rainfall. One of the problem when dealing with debris flow susceptible areas is that the eroded surface is usually very shallow and it can be masked by the vegetation already few years after a landslide event. Therefore, debris flow prone areas very often suffer from lack of reliable landslide inventories necessary to calibrate and validate susceptibility models. In order to deal with this problem, transferability procedure (spatial partition) have already proved to be efficient in areas which show analogous topographic, lithological and climatic characteristics. A procedure to evaluate whether it is possible to apply model transferability is here proposed. This approach is based on the assumption that debris flow trigger in different locations under similar topographic conditions and includes: i) a test of similarity between training and test areas aimed at identifying thresholds in catchment similarity which allow to successfully perform the transferability; ii) the calibration of the susceptibility model in the training area; iii) the validation of the model on the test area. The debris flow susceptibility is here evaluated using a stochastic approach and the all procedure is implemented in a R script which can be easily used to test the procedure in other catchments. The study areas chosen to perform this study are located in the Messina province (southern Italy) respectively on the Ionian sector (Itala catchment) and on the Tyrrhenian sector (Saponara catchment). Itala catchment was hit by the sadly known debris flow event of the 1st October 2009 (37 fatalities and huge damages) while Saponara catchment on the 22nd November 2011 (only two years after the 2009 event) experienced a very

  14. Calibration of a Numerical Model for Heat Transfer and Fluid Flow in an Extruder

    DEFF Research Database (Denmark)

    Hofstätter, Thomas; Pedersen, David Bue; Nielsen, Jakob Skov;

    2016-01-01

    This paper discusses experiments performed in order to validate simulations on a fused deposition modelling (FDM) extruder. The nozzle has been simulated in terms of heat transfer and fluid flow. In order to calibrate and validate these simulations, experiments were performed giving a significant...... look into the physical behaviour of the nozzle, heating and cooling systems. Experiments on the model were performed at different sub-mm diameters of the extruder. Physical parameters of the model – especially temperature dependent parameters – were set into analytical relationships in order to receive...... dynamical parameters. This research sets the foundation for further research within melted extrusion based additive manufacturing. The heating process of the extruder will be described and a note on the material feeding will be given....

  15. Enhanced Single Seed Trait Predictions in Soybean (Glycine max) and Robust Calibration Model Transfer with Near-Infrared Reflectance Spectroscopy.

    Science.gov (United States)

    Hacisalihoglu, Gokhan; Gustin, Jeffery L; Louisma, Jean; Armstrong, Paul; Peter, Gary F; Walker, Alejandro R; Settles, A Mark

    2016-02-10

    Single seed near-infrared reflectance (NIR) spectroscopy predicts soybean (Glycine max) seed quality traits of moisture, oil, and protein. We tested the accuracy of transferring calibrations between different single seed NIR analyzers of the same design by collecting NIR spectra and analytical trait data for globally diverse soybean germplasm. X-ray microcomputed tomography (μCT) was used to collect seed density and shape traits to enhance the number of soybean traits that can be predicted from single seed NIR. Partial least-squares (PLS) regression gave accurate predictive models for oil, weight, volume, protein, and maximal cross-sectional area of the seed. PLS models for width, length, and density were not predictive. Although principal component analysis (PCA) of the NIR spectra showed that black seed coat color had significant signal, excluding black seeds from the calibrations did not impact model accuracies. Calibrations for oil and protein developed in this study as well as earlier calibrations for a separate NIR analyzer of the same design were used to test the ability to transfer PLS regressions between platforms. PLS models built from data collected on one NIR analyzer had minimal differences in accuracy when applied to spectra collected from a sister device. Model transfer was more robust when spectra were trimmed from 910 to 1679 nm to 955-1635 nm due to divergence of edge wavelengths between the two devices. The ability to transfer calibrations between similar single seed NIR spectrometers facilitates broader adoption of this high-throughput, nondestructive, seed phenotyping technology.

  16. FDTD Simulation of Transfer Impedance of Calibration Current Target for ESD Testing

    Science.gov (United States)

    Gotoh, Tsuyoshi; Taka, Yoshinori; Mori, Ikuko; Fujiwara, Osamu; Ishigami, Shinobu; Yamanaka, Yukio

    Transient electromagnetic fields caused by electrostatic discharges (ESDs) are known to cause a serious failure in high-tech information devices. From this perspective, an international standard for ESD testing or IEC 61000-4-2 has been specified, in which a detailed waveform of the discharge current to be injected onto equipment under test (EUT) by an ESD-gun is prescribed, and is also required for contact discharges onto a calibration current target called Pellegrini target, which is specially designed so that the waveform of an injected current and its observed voltage coincides. However, the coincidence relationship should not hold at high frequencies due to parasitic components which the target has. In this study, we constructed an FDTD model of the Pellegrini target, and calculated its transfer impedance by simulating the injected current onto the target and the corresponding output voltage in combination with an FDTD model being previously developed for an ESD-gun. As a result, we found that the target has a transfer impedance of almost 1-j0 Ω at less than a few hundred MHz, though it has resonance phenomena around 2-3 GHz, which was validated by measuring scattering parameters of the target through a transmission line with a network analyzer. Using the FDTD-calculated transfer impedance, we reconstructed injected current waveforms for contact and air discharges of an ESD-gun from their observed voltages to reveal that the injected current waveform is almost identical to the observed voltage waveform for the contact discharge, while the injected currents for the air discharges have smaller peaks and longer rise times than the observed voltages do due to the presence of resonance properties of the transfer impedance.

  17. Nitrous oxide emissions from cropland: a procedure for calibrating the DayCent biogeochemical model using inverse modelling

    Science.gov (United States)

    Rafique, Rashad; Fienen, Michael N.; Parkin, Timothy B.; Anex, Robert P.

    2013-01-01

    DayCent is a biogeochemical model of intermediate complexity widely used to simulate greenhouse gases (GHG), soil organic carbon and nutrients in crop, grassland, forest and savannah ecosystems. Although this model has been applied to a wide range of ecosystems, it is still typically parameterized through a traditional “trial and error” approach and has not been calibrated using statistical inverse modelling (i.e. algorithmic parameter estimation). The aim of this study is to establish and demonstrate a procedure for calibration of DayCent to improve estimation of GHG emissions. We coupled DayCent with the parameter estimation (PEST) software for inverse modelling. The PEST software can be used for calibration through regularized inversion as well as model sensitivity and uncertainty analysis. The DayCent model was analysed and calibrated using N2O flux data collected over 2 years at the Iowa State University Agronomy and Agricultural Engineering Research Farms, Boone, IA. Crop year 2003 data were used for model calibration and 2004 data were used for validation. The optimization of DayCent model parameters using PEST significantly reduced model residuals relative to the default DayCent parameter values. Parameter estimation improved the model performance by reducing the sum of weighted squared residual difference between measured and modelled outputs by up to 67 %. For the calibration period, simulation with the default model parameter values underestimated mean daily N2O flux by 98 %. After parameter estimation, the model underestimated the mean daily fluxes by 35 %. During the validation period, the calibrated model reduced sum of weighted squared residuals by 20 % relative to the default simulation. Sensitivity analysis performed provides important insights into the model structure providing guidance for model improvement.

  18. Numerical modeling of two-dimensional heat-transfer and temperature-based calibration using simulated annealing optimization method: Application to gas metal arc welding

    Directory of Open Access Journals (Sweden)

    Bjelić Mišo B.

    2016-01-01

    Full Text Available Simulation models of welding processes allow us to predict influence of welding parameters on the temperature field during welding and by means of temperature field and the influence to the weld geometry and microstructure. This article presents a numerical, finite-difference based model of heat transfer during welding of thin sheets. Unfortunately, accuracy of the model depends on many parameters, which cannot be accurately prescribed. In order to solve this problem, we have used simulated annealing optimization method in combination with presented numerical model. This way, we were able to determine uncertain values of heat source parameters, arc efficiency, emissivity and enhanced conductivity. The calibration procedure was made using thermocouple measurements of temperatures during welding for P355GH steel. The obtained results were used as input for simulation run. The results of simulation showed that represented calibration procedure could significantly improve reliability of heat transfer model. [National CEEPUS Office of Czech Republic (project CIII-HR-0108-07-1314 and to the Ministry of Education and Science of the Republic of Serbia (project TR37020

  19. Enhanced single seed trait predictions in soybean (Glycine max) and robust calibration model transfer with near infrared reflectance spectroscopy

    Science.gov (United States)

    Single seed near infrared reflectance (NIR) spectroscopy predicts soybean (Glycine max) seed quality traits of moisture, oil, and protein. We tested the accuracy of transferring calibrations between different single seed NIR analyzers of the same design by collecting NIR spectra and analytical trait...

  20. NASA Glenn Icing Research Tunnel: 2014 and 2015 Cloud Calibration Procedures and Results

    Science.gov (United States)

    Steen, Laura E.; Ide, Robert F.; Van Zante, Judith F.; Acosta, Waldo J.

    2015-01-01

    This report summarizes the current status of the NASA Glenn Research Center (GRC) Icing Research Tunnel cloud calibration: specifically, the cloud uniformity, liquid water content, and drop-size calibration results from both the January-February 2014 full cloud calibration and the January 2015 interim cloud calibration. Some aspects of the cloud have remained the same as what was reported for the 2014 full calibration, including the cloud uniformity from the Standard nozzles, the drop-size equations for Standard and Mod1 nozzles, and the liquid water content for large-drop conditions. Overall, the tests performed in January 2015 showed good repeatability to 2014, but there is new information to report as well. There have been minor updates to the Mod1 cloud uniformity on the north side of the test section. Also, successful testing with the OAP-230Y has allowed the IRT to re-expand its operating envelopes for large-drop conditions to a maximum median volumetric diameter of 270 microns. Lastly, improvements to the collection-efficiency correction for the SEA multi-wire have resulted in new calibration equations for Standard- and Mod1-nozzle liquid water content.

  1. Calibration procedure for a potato crop growth model using information from across Europe

    DEFF Research Database (Denmark)

    Heidmann, Tove; Tofteng, C; Abrahamsen, P;

    2008-01-01

    to simulate the field experiments. An initial potato parameterisation was generated from an independent dataset and was used for site-specific calibrations. At those sites where the same variety was used for all 3 years, the calibration of the initial potato model was based on the first 2 years using the last......In the FertOrgaNic EU project, 3 years of field experiments with drip irrigation and fertigation were carried out at six different sites across Europe, involving seven different varieties of potato. The Daisy model, which simulates plant growth together with water and nitrogen dynamics, was used...

  2. Binary Pseudo-Random Gratings and Arrays for Calibration of Modulation Transfer Functions of Surface Profilometers

    Energy Technology Data Exchange (ETDEWEB)

    Barber, Samuel K.; Anderson, Erik D.; Cambie, Rossana; McKinney, Wayne R.; Takacs, Peter Z.; Stover, John C.; Voronov, Dmitriy L.; Yashchuk, Valeriy V.

    2009-09-11

    A technique for precise measurement of the modulation transfer function (MTF), suitable for characterization of a broad class of surface profilometers, is investigated in detail. The technique suggested in [Proc. SPIE 7077-7, (2007), Opt. Eng. 47(7), 073602-1-5 (2008)]is based on use of binary pseudo-random (BPR) gratings and arrays as standard MTF test surfaces. Unlike most conventional test surfaces, BPR gratings and arrays possess white-noise-like inherent power spectral densities (PSD), allowing the direct determination of the one- and two-dimensional MTF, respectively, with a sensitivity uniform over the entire spatial frequency range of a profiler. In the cited work, a one dimensional realization of the suggested method based on use of BPR gratings has been demonstrated. Here, a high-confidence of the MTF calibration technique is demonstrated via cross comparison measurements of a number of two dimensional BPR arrays using two different interferometric microscopes and a scatterometer. We also present the results of application of the experimentally determined MTF correction to the measurement taken with the MicromapTM-570 interferometric microscope of the surface roughness of a super-polished test mirror. In this particular case, without accounting for the instrumental MTF, the surface rms roughness over half of the instrumental spatial frequency bandwidth would be underestimated by a factor of approximately 1.4.

  3. Randomized trial to examine procedure-to-procedure transfer in laparoscopic simulator training

    DEFF Research Database (Denmark)

    Bjerrum, F; Sorensen, J L; Konge, L

    2016-01-01

    -centre educational superiority trial. Surgical novices practised basic skills on a laparoscopic virtual reality simulator. On reaching proficiency, participants were randomized to proficiency-based training. The intervention group practised two procedures on the simulator (appendicectomy followed by salpingectomy...

  4. 40 CFR 90.424 - Dilute sampling procedures-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... to the National Institute for Standards and Testing (NIST) and serves as the reference value (NIST “true” value) for the CVS calibration. (Note: In no case should an upstream screen or other restriction... has been obtained, verification of the entire system may be performed by injecting a known mass of...

  5. Acute infection with Propionibacterium acnes after a Latarjet coracoid transfer procedure: a case report.

    Science.gov (United States)

    Erickson, Brandon J; Bhatia, Sanjeev; Biswas, Debdut; Verma, Nikhil N

    2016-06-01

    Coracoid transfer procedures are highly effective at improving glenohumeral stability in patients with recurrent shoulder instability; complication rates from this procedure are generally low. We present the first case in the literature of a Propionibacterium acnes (P. acnes) infection following a coracoid transfer. The case stresses the importance of proper antibiotic prophylaxis for patients undergoing Latarjet procedures, as well as the workup for a painful postoperative shoulder, and the need to maintain cultures from the shoulder for up to 21 days to isolate P. acnes.

  6. Boiling heat transfer on fins – experimental and numerical procedure

    Directory of Open Access Journals (Sweden)

    Orzechowski T.

    2014-03-01

    Full Text Available The paper presents the research methodology, the test facility and the results of investigations into non-isothermal surfaces in water boiling at atmospheric pressure, together with a discussion of errors. The investigations were conducted for two aluminium samples with technically smooth surfaces and thickness of 4 mm and 10 mm, respectively. For the sample of lower thickness, on the basis of the surface temperature distribution measured with an infrared camera, the local heat flux and the heat transfer coefficient were determined and shown in the form of a boiling curve. For the thicker sample, for which 1-D model cannot be used, numerical calculations were conducted. They resulted in obtaining the values of the local heat flux on the surface the invisible to the infrared, camera i.e. on the side on which the boiling of the medium proceeds.

  7. Nonlinear calibration transfer based on hierarchical Bayesian models and Lagrange Multipliers: Error bounds of estimates via Monte Carlo - Markov Chain sampling.

    Science.gov (United States)

    Seichter, Felicia; Vogt, Josef; Radermacher, Peter; Mizaikoff, Boris

    2017-01-25

    The calibration of analytical systems is time-consuming and the effort for daily calibration routines should therefore be minimized, while maintaining the analytical accuracy and precision. The 'calibration transfer' approach proposes to combine calibration data already recorded with actual calibrations measurements. However, this strategy was developed for the multivariate, linear analysis of spectroscopic data, and thus, cannot be applied to sensors with a single response channel and/or a non-linear relationship between signal and desired analytical concentration. To fill this gap for a non-linear calibration equation, we assume that the coefficients for the equation, collected over several calibration runs, are normally distributed. Considering that coefficients of an actual calibration are a sample of this distribution, only a few standards are needed for a complete calibration data set. The resulting calibration transfer approach is demonstrated for a fluorescence oxygen sensor and implemented as a hierarchical Bayesian model, combined with a Lagrange Multipliers technique and Monte-Carlo Markov-Chain sampling. The latter provides realistic estimates for coefficients and prediction together with accurate error bounds by simulating known measurement errors and system fluctuations. Performance criteria for validation and optimal selection of a reduced set of calibration samples were developed and lead to a setup which maintains the analytical performance of a full calibration. Strategies for a rapid determination of problems occurring in a daily calibration routine, are proposed, thereby opening the possibility of correcting the problem just in time.

  8. Validation Tests of Open-Source Procedures for Digital Camera Calibration and 3d Image-Based Modelling

    Science.gov (United States)

    Toschi, I.; Rivola, R.; Bertacchini, E.; Castagnetti, C.; Dubbini, M.; Capra, A.

    2013-07-01

    Among the many open-source software solutions recently developed for the extraction of point clouds from a set of un-oriented images, the photogrammetric tools Apero and MicMac (IGN, Institut Géographique National) aim to distinguish themselves by focusing on the accuracy and the metric content of the final result. This paper firstly aims at assessing the accuracy of the simplified and automated calibration procedure offered by the IGN tools. Results obtained with this procedure were compared with those achieved with a test-range calibration approach using a pre-surveyed laboratory test-field. Both direct and a-posteriori validation tests turned out successfully showing the stability and the metric accuracy of the process, even when low textured or reflective surfaces are present in the 3D scene. Afterwards, the possibility of achieving accurate 3D models from the subsequently extracted dense point clouds is also evaluated. Three different types of sculptural elements were chosen as test-objects and "ground-truth" data were acquired with triangulation laser scanners. 3D models derived from point clouds oriented with a simplified relative procedure show a suitable metric accuracy: all comparisons delivered a standard deviation of millimeter-level. The use of Ground Control Points in the orientation phase did not improve significantly the accuracy of the final 3D model, when a small figure-like corbel was used as test-object.

  9. Building Energy Simulation Test for Existing Homes (BESTEST-EX): Instructions for Implementing the Test Procedure, Calibration Test Reference Results, and Example Acceptance-Range Criteria

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, R.; Polly, B.; Bianchi, M.; Neymark, J.; Kennedy, M.

    2011-08-01

    This publication summarizes building energy simulation test for existing homes (BESTEST-EX): instructions for implementing the test procedure, calibration tests reference results, and example acceptance-range criteria.

  10. Characterizing Thermal features in Norris Basin, Yellowstone National Park, Using Multi- spectral Remote Sensing Data and Dynamic Calibration Procedures

    Science.gov (United States)

    Hardy, C. C.; Queen, L. P.; Heasler, H. P.; Jaworowski, C.

    2007-12-01

    A thermal infrared remote sensing project was implemented to develop methods for identifying, classifying, and mapping thermal features. This study is directed at geothermal features, with the expectation that new protocols developed here will apply to the wildland fire thermal environment. Airborne multi-spectral digital imagery were acquired over the geothermally active Norris Basin region of Yellowstone National Park, USA. Two image acquisitions were flown, with one acquisition near solar noon and the other at night. Raw data from the five sensors were uncalibrated, so a vicarious calibration procedure was developed to compute reflectance for the visible and NIR bands using an independently calibrated hyperspectral dataset. Calibration of the thermal sensor band utilized a dynamic, in-scene calibration procedure that exploited natural, pseudo-invariant thermal reference targets instrumented with in situ kinetic temperature recorders. The calibrated reflectance and radiant temperature data from each acquisition were processed and analyzed to develop a suite of thermal attributes, including radiant temperatures, a daytime-nighttime temperature difference (DeltaT), albedo, an albedo derivative (one minus albedo), and apparent thermal inertia (ATI). The albedo terms were computed using a published weighed-average albedo algorithm based on ratios of the narrowband red and near-infrared (NIR) reflectances to total solar irradiance for the respective red and NIR bandpasses. The weighing factors for each band were the proportion of total solar irradiance incident on the surface within each segment represented by a respective bandpass. In the absence of verifiable "truth," a step-wise chain of unsupervised classification and multivariate analysis exercises was performed, drawing heavily on "fuzzy truth" to assess the quality, efficiency, and efficacy of classification procedures and results. A final classification synthesizes a "geothermal phenomenology" comprised of

  11. Procedures for the relative calibration of the SiPM gain on ASTRI SST-2M camera

    Science.gov (United States)

    Impiombato, D.; Catalano, O.; Giarrusso, S.; Mineo, T.; Rosa, G. La; Gargano, C.; Sangiorgi, P.; Segreto, A.; Sottile, G.; Bonanno, G.; Garozzo, S.; Grillo, A.; Marano, D.; Romeo, G.; Gimenes, R.

    2017-02-01

    ASTRI SST-2M is one of the prototypes of the small size class of telescopes proposed for the Cherenkov Telescope Array. Its optical design is based on a dual-mirror Schwarzschild-Couder configuration, and the camera is composed by a matrix of monolithic multipixel silicon photomultipliers managed by ad-hoc tailored front-end electronics. This paper describes the procedures for the gain calibration on the ASTRI SST-2M. Since the SiPM gain depends on the operative voltage and the temperature, we adjust the operative voltages for all sensors to have equal gains at a reference temperature. We then correct gain variations caused by temperature changes by adjusting the operating voltage of each sensor. For that purpose the SiPM gain dependence on operating voltage and on temperature have been measured. In addition, we present the calibration procedures and the results of the experimental measurements to evaluate, for each pixel, the parameters necessary to make the trigger uniform over the whole focal plane.

  12. Community Radiative Transfer Model for Inter-Satellites Calibration and Verification

    Science.gov (United States)

    Liu, Q.; Nalli, N. R.; Ignatov, A.; Garrett, K.; Chen, Y.; Weng, F.; Boukabara, S. A.; van Delst, P. F.; Groff, D. N.; Collard, A.; Joseph, E.; Morris, V. R.; Minnett, P. J.

    2014-12-01

    Developed at the Joint Center for Satellite Data Assimilation, the Community Radiative Transfer Model (CRTM) [1], operationally supports satellite radiance assimilation for weather forecasting. The CRTM also supports JPSS/NPP and GOES-R missions [2] for instrument calibration, validation, monitoring long-term trending, and satellite retrieved products [3]. The CRTM is used daily at the NOAA NCEP to quantify the biases and standard deviations between radiance simulations and satellite radiance measurements in a time series and angular dependency. The purposes of monitoring the data assimilation system are to ensure the proper performance of the assimilation system and to diagnose problems with the system for future improvements. The CRTM is a very useful tool for cross-sensor verifications. Using the double difference method, it can remove the biases caused by slight differences in spectral response and geometric angles between measurements of the two instruments. The CRTM is particularly useful to reduce the difference between instruments for climate studies [4]. In this study, we will carry out the assessment of the Suomi National Polar-orbiting Partnership (SNPP) [5] Cross-track Infrared Sounder (CrIS) data [6], Advanced Technology Microwave Sounder (ATMS) data, and data for Visible Infrared Imaging Radiometer Suite (VIIRS) [7][8] thermal emissive bands. We use dedicated radiosondes and surface data acquired from NOAA Aerosols and Ocean Science Expeditions (AEROSE) [9]. The high quality radiosondes were launched when Suomi NPP flew over NOAA Ship Ronald H. Brown situated in the tropical Atlantic Ocean. The atmospheric data include profiles of temperature, water vapor, and ozone, as well as total aerosol optical depths. The surface data includes air temperature and humidity at 2 meters, skin temperature (Marine Atmospheric Emitted Radiance Interferometer, M-AERI [10]), surface temperature, and surface wind vector. [1] Liu, Q., and F. Weng, 2006: JAS [2] Liu, Q

  13. Instructional strategy effects on the retention and transfer of procedures of different difficulty level

    NARCIS (Netherlands)

    Jelsma, Otto; Pieters, Jules M.

    1989-01-01

    In the present study, the effects of two instructional strategies on the retention and transfer of procedures of different difficulty level were investigated. Difficulty level was manipulated by providing a different number of cues during training. The instructional strategies differed with respect

  14. Finite element procedures for coupled linear analysis of heat transfer, fluid and solid mechanics

    Science.gov (United States)

    Sutjahjo, Edhi; Chamis, Christos C.

    1993-01-01

    Coupled finite element formulations for fluid mechanics, heat transfer, and solid mechanics are derived from the conservation laws for energy, mass, and momentum. To model the physics of interactions among the participating disciplines, the linearized equations are coupled by combining domain and boundary coupling procedures. Iterative numerical solution strategy is presented to solve the equations, with the partitioning of temporal discretization implemented.

  15. Proceduralization, transfer of training and retention of knowledge as a result of output practice

    Directory of Open Access Journals (Sweden)

    Mehrnoosh Fakharzadeh

    2013-11-01

    Full Text Available This study investigated the effect of output practice on the proceduralization, transfer and retention of knowledge on English modals, adopting Anderson's ACT-R model of skill acquisition. A pretest posttest and delayed posttest design was used where the procedural knowledge on production skill was specifically operationalized through the groups’ performance on Dual Task Timed Completion Test and transfer of training was measures through a Dual Task Timed Grammaticality Judgment Test. Two intact classes of intermediate EFL learners were randomly assigned to treatment and control groups. The output group (n= 27 received explicit grammar instruction, a combination of three output tasks including dictogloss, individual text reconstruction, and corrected-close translation, and feedback. The control group (n=25 were just exposed to the identical texts trough listening and reading tasks followed by some questions irrelevant to the target structure. Results showed that on the posttest, three days after the last treatment session, the output group outperformed the control group in both measures of procedural knowledge and transfer of knowledge. As for retention, forty days later, the output group’s performance was still significantly better than that on the pretest. This group also succeeded in retaining its outperformance on both measures of procedural knowledge and transfer of knowledge delayed posttest. The results may help language teachers design more effective activities for the learners considering the institutional constraints.

  16. For geological investigations with airborne thermal infrared multispectral images: Transfer of calibration from laboratory spectrometer to TIMS as alternative for removing atmospheric effects

    Science.gov (United States)

    Edgett, Kenneth S.; Anderson, Donald L.

    1995-01-01

    This paper describes an empirical method to correct TIMS (Thermal Infrared Multispectral Scanner) data for atmospheric effects by transferring calibration from a laboratory thermal emission spectrometer to the TIMS multispectral image. The method does so by comparing the laboratory spectra of samples gathered in the field with TIMS 6-point spectra for pixels at the location of field sampling sites. The transference of calibration also makes it possible to use spectra from the laboratory as endmembers in unmixing studies of TIMS data.

  17. Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Y.; Wuu, Cheng-Shie; Maryanski, Marek J. [Department of Radiation Oncology, Columbia University, New York, New York 10032 (United States); Department of Radiation Oncology, Columbia University, New York, New York 10032 and MGS Research Inc., Madison, Connecticut 06443 (United States)

    2010-02-15

    The dose response of the BANG 3 polymer gel dosimeter (MGS Research Inc., Madison, CT) was studied using the OCTOPUS laser CT scanner (MGS Research Inc., Madison, CT). Six 17 cm diameter and 12 cm high Barex cylinders, and 18 small glass vials were used to house the gel. The gel phantoms were irradiated with 6 and 10 MV photons, as well as 12 and 16 MeV electrons using a Varian Clinac 2100EX. Three calibration methods were used to obtain the dose response curves: (a) Optical density measurements on the 18 glass vials irradiated with graded doses from 0 to 4 Gy using 6 or 10 MV large field irradiations; (b) optical-CT scanning of Barex cylinders irradiated with graded doses (0.5, 1, 1.5, and 2 Gy) from four adjacent 4x4 cm{sup 2} photon fields or 6x6 cm{sup 2} electron fields; and (c) percent depth dose (PDD) comparison of optical-CT scans with ion chamber measurements for 6x6 cm{sup 2}, 12 and 16 MeV electron fields. The dose response of the BANG 3 gel was found to be linear and energy independent within the uncertainties of the experimental methods (about 3%). The slopes of the linearly fitted dose response curves (dose sensitivities) from the four field irradiations (0.0752{+-}3%, 0.0756{+-}3%, 0.0767{+-}3%, and 0.0759{+-}3% cm{sup -1} Gy{sup -1}) and the PDD matching methods (0.0768{+-}3% and 0.0761{+-}3% cm{sup -1} Gy{sup -1}) agree within 2.2%, indicating a good reproducibility of the gel dose response within phantoms of the same geometry. The dose sensitivities from the glass vial approach are different from those of the cylindrical Barex phantoms by more than 30%, owing probably to the difference in temperature inside the two types of phantoms during gel formation and irradiation, and possible oxygen contamination of the glass vial walls. The dose response curve obtained from the PDD matching approach with 16 MeV electron field was used to calibrate the gel phantom irradiated with the 12 MeV, 6x6 cm{sup 2} electron field. Three-dimensional dose distributions

  18. REMINDER Saved Leave Scheme (SLS) : Simplified procedure for the transfer of leave to saved leave accounts

    CERN Document Server

    HR Division

    2001-01-01

    As part of the process of streamlining procedures, the HR and AS Divisions have jointly developed a system whereby annual and compensatory leave will henceforth be automatically transferred1) to saved leave accounts. Under the provisions of the voluntary saved leave scheme (SLS), a maximum total of 10 days'2)Previously, every person taking part in the scheme has been individually issued with a form for the purposes of requesting the transfer of leave to the leave account and the transfer has then had to be done manually by HR Division. To streamline the procedure, unused leave of all those taking part in the saved leave scheme at the closure of the leave-year accounts will henceforth be transferred automatically to the saved leave account on that date. This simplification is in the interest of all parties concerned. This automatic transfer procedure has a number of advantages for participants in the SLS scheme. First, staff members will no longer have to take any administrative steps. Secondly, the new proced...

  19. Life History Traits and Niche Instability Impact Accuracy and Temporal Transferability for Historically Calibrated Distribution Models of North American Birds.

    Science.gov (United States)

    Wogan, Guinevere O U

    2016-01-01

    A primary assumption of environmental niche models (ENMs) is that models are both accurate and transferable across geography or time; however, recent work has shown that models may be accurate but not highly transferable. While some of this is due to modeling technique, individual species ecologies may also underlie this phenomenon. Life history traits certainly influence the accuracy of predictive ENMs, but their impact on model transferability is less understood. This study investigated how life history traits influence the predictive accuracy and transferability of ENMs using historically calibrated models for birds. In this study I used historical occurrence and climate data (1950-1990s) to build models for a sample of birds, and then projected them forward to the 'future' (1960-1990s). The models were then validated against models generated from occurrence data at that 'future' time. Internal and external validation metrics, as well as metrics assessing transferability, and Generalized Linear Models were used to identify life history traits that were significant predictors of accuracy and transferability. This study found that the predictive ability of ENMs differs with regard to life history characteristics such as range, migration, and habitat, and that the rarity versus commonness of a species affects the predicted stability and overlap and hence the transferability of projected models. Projected ENMs with both high accuracy and transferability scores, still sometimes suffered from over- or under- predicted species ranges. Life history traits certainly influenced the accuracy of predictive ENMs for birds, but while aspects of geographic range impact model transferability, the mechanisms underlying this are less understood.

  20. Procedural learning, consolidation, and transfer of a new skill in Developmental Coordination Disorder.

    Science.gov (United States)

    Lejeune, Caroline; Wansard, Murielle; Geurten, Marie; Meulemans, Thierry

    2016-01-01

    The aim of this study was to explore the differences in procedural learning abilities between children with DCD and typically developing children by investigating the steps that lead to skill automatization (i.e., the stages of fast learning, consolidation, and slow learning). Transfer of the skill to a new situation was also assessed. We tested 34 children aged 6-12 years with and without DCD on a perceptuomotor adaptation task, a form of procedural learning that is thought to involve the cerebellum and the basal ganglia (regions whose impairment has been associated with DCD) but also other brain areas including frontal regions. The results showed similar rates of learning, consolidation, and transfer in DCD and control children. However, the DCD children's performance remained slower than that of controls throughout the procedural task and they reached a lower asymptotic performance level; the difficulties observed at the outset did not diminish with practice.

  1. Outcome-Based Comparison of Percutaneous Procedures for Urinary Lithiasis with Calibre of Instrumentation less than 12Fr.

    Science.gov (United States)

    Wells, H; Rukin, N; Wright, A; Somani, B K

    2015-08-01

    Renal stone disease is becoming increasingly prevalent globally. With a rise in stone disease worldwide, there is also a relative increase in the rates of surgical intervention. Technological advances have allowed a move towards minimising the complications rates and length of stay with a reduction in invasiveness and size of instruments. A trend for minimising the percutaneous tract size has been noted in percutaneous nephrolithotomy (PCNL) for renal stones. The management has shifted from open surgery to standard PCNL, mini-PCNL and the latest ultra-mini and micro-PCNL techniques. There is a need to compare outcomes for the ever-advancing technologies, such as the smaller calibre of instruments, to assess risk-benefit in practice. This review looks at outcome-based comparison of percutaneous procedures for urinary lithiasis with instruments <12Fr in size.

  2. NASA Glenn Icing Research Tunnel: Upgrade and Cloud Calibration

    Science.gov (United States)

    VanZante, Judith Foss; Ide, Robert F.; Steen, Laura E.

    2012-01-01

    In 2011, NASA Glenn s Icing Research Tunnel underwent a major modification to it s refrigeration plant and heat exchanger. This paper presents the results of the subsequent full cloud calibration. Details of the calibration procedure and results are presented herein. The steps include developing a nozzle transfer map, establishing a uniform cloud, conducting a drop sizing calibration and finally a liquid water content calibration. The goal of the calibration is to develop a uniform cloud, and to build a transfer map from the inputs of air speed, spray bar atomizing air pressure and water pressure to the output of median volumetric droplet diameter and liquid water content.

  3. An Innovative Procedure for Calibration of Strapdown Electro-Optical Sensors Onboard Unmanned Air Vehicles

    Directory of Open Access Journals (Sweden)

    Attilio Rispoli

    2010-01-01

    Full Text Available This paper presents an innovative method for estimating the attitude of airborne electro-optical cameras with respect to the onboard autonomous navigation unit. The procedure is based on the use of attitude measurements under static conditions taken by an inertial unit and carrier-phase differential Global Positioning System to obtain accurate camera position estimates in the aircraft body reference frame, while image analysis allows line-of-sight unit vectors in the camera based reference frame to be computed. The method has been applied to the alignment of the visible and infrared cameras installed onboard the experimental aircraft of the Italian Aerospace Research Center and adopted for in-flight obstacle detection and collision avoidance. Results show an angular uncertainty on the order of 0.1° (rms.

  4. An innovative procedure for calibration of strapdown electro-optical sensors onboard unmanned air vehicles.

    Science.gov (United States)

    Fasano, Giancarmine; Accardo, Domenico; Moccia, Antonio; Rispoli, Attilio

    2010-01-01

    This paper presents an innovative method for estimating the attitude of airborne electro-optical cameras with respect to the onboard autonomous navigation unit. The procedure is based on the use of attitude measurements under static conditions taken by an inertial unit and carrier-phase differential Global Positioning System to obtain accurate camera position estimates in the aircraft body reference frame, while image analysis allows line-of-sight unit vectors in the camera based reference frame to be computed. The method has been applied to the alignment of the visible and infrared cameras installed onboard the experimental aircraft of the Italian Aerospace Research Center and adopted for in-flight obstacle detection and collision avoidance. Results show an angular uncertainty on the order of 0.1° (rms).

  5. Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines

    NARCIS (Netherlands)

    Hekkenberg, R.T.; Richards, A.; Beissner, K.; Zeqiri, B.; Bezemer, R.A.; Hodnett, M.; Prout, G.; Cantrall, C.

    2006-01-01

    Ultrasound (US) physiotherapy as a clinical treatment is extremely common in the Western world. Internationally, regulation to ensure safe application of US physiotherapy by regular calibration ranges from nil to mandatory. The need for a portable power standard (PPS) has been addressed within a Eur

  6. Saved Leave Scheme (SLS) : Simplified procedure for the transfer of leave to saved leave accounts

    CERN Multimedia

    HR Division

    2001-01-01

    As part of the process of streamlining procedures, the HR and AS Divisions have jointly developed a system whereby annual and compensatory leave will henceforth be automatically transferred1) to saved leave accounts. Under the provisions of the voluntary saved leave scheme (SLS), a maximum total of 10 days'2) annual and compensatory leave (excluding saved leave accumulated in accordance with the provisions of Administrative Circular No. 22 B) can be transferred to the saved leave account at the end of the leave year (30 September). Previously, every person taking part in the scheme has been individually issued with a form for the purposes of requesting the transfer of leave to the leave account and the transfer has then had to be done manually by HR Division. To streamline the procedure, unused leave of all those taking part in the saved leave scheme at the closure of of the leave-year accounts will henceforth be transferred automatically to the saved leave account on that date. This simplification is in the ...

  7. A robust calibration technique for acoustic emission systems based on momentum transfer from a ball drop

    Science.gov (United States)

    McLaskey, Gregory C.; Lockner, David A.; Kilgore, Brian D.; Beeler, Nicholas M.

    2015-01-01

    We describe a technique to estimate the seismic moment of acoustic emissions and other extremely small seismic events. Unlike previous calibration techniques, it does not require modeling of the wave propagation, sensor response, or signal conditioning. Rather, this technique calibrates the recording system as a whole and uses a ball impact as a reference source or empirical Green’s function. To correctly apply this technique, we develop mathematical expressions that link the seismic moment $M_{0}$ of internal seismic sources (i.e., earthquakes and acoustic emissions) to the impulse, or change in momentum $\\Delta p $, of externally applied seismic sources (i.e., meteor impacts or, in this case, ball impact). We find that, at low frequencies, moment and impulse are linked by a constant, which we call the force‐moment‐rate scale factor $C_{F\\dot{M}} = M_{0}/\\Delta p$. This constant is equal to twice the speed of sound in the material from which the seismic sources were generated. Next, we demonstrate the calibration technique on two different experimental rock mechanics facilities. The first example is a saw‐cut cylindrical granite sample that is loaded in a triaxial apparatus at 40 MPa confining pressure. The second example is a 2 m long fault cut in a granite sample and deformed in a large biaxial apparatus at lower stress levels. Using the empirical calibration technique, we are able to determine absolute source parameters including the seismic moment, corner frequency, stress drop, and radiated energy of these magnitude −2.5 to −7 seismic events.

  8. Transient, compressible heat and mass transfer in porous media using the strongly implicit iteration procedure.

    Science.gov (United States)

    Curry, D. M.; Cox, J. E.

    1972-01-01

    Coupled nonlinear partial differential equations describing heat and mass transfer in a porous matrix are solved in finite difference form with the aid of a new iterative technique (the strongly implicit procedure). Example numerical results demonstrate the characteristics of heat and mass transport in a porous matrix such as a charring ablator. It is emphasized that multidimensional flow must be considered when predicting the thermal response of a porous material subjected to nonuniform boundary conditions.

  9. Clinic based transfer of the N{sub D,w}{sup 6{sup 0Co}} calibration coefficient using a linear accelerator

    Energy Technology Data Exchange (ETDEWEB)

    Abdel-Rahman, W.; Evans, M. D. C.; Serre, L.; McCaffrey, J. P.; Podgorsak, E. B.; Seuntjens, J. P. [Department of Medical Physics, McGill University Health Centre, 1650 Avenue Cedar, Montreal, Quebec H3G 1A4 (Canada); Windsor Regional Cancer Center, 2220 Kildre Road, Windsor, Ontario N8W 2X3 (Canada); Ionizing Radiation Standards, National Research Council of Canada, Ottawa, Ontario K1A 0R6 (Canada); Department of Medical Physics, McGill University Health Centre, 1650 Avenue Cedar, Montreal, Quebec H3G 1A4 (Canada)

    2009-03-15

    Ionization chambers used for reference dosimetry require a local secondary standard ionization chamber with a {sup 60}Co absorbed dose to water calibration coefficient N{sub D,w}{sup 6{sup 0Co}} traceable to a national primary standards dosimetry laboratory or an accredited secondary dosimetry calibration laboratory. Clinic based (in-house) transfer of this coefficient to tertiary reference ionization chambers has traditionally been accomplished with chamber cross calibration in water using a {sup 60}Co beam; however, access to {sup 60}Co teletherapy machines has become increasingly limited for clinic based physicists. In this work, the accuracy of alternative methods of transferring the N{sub D,w}{sup 6{sup 0Co}} calibration coefficient using 6 and 18 MV photon beams from a linear accelerator in lieu of {sup 60}Co has been investigated for five different setups and four commonly used chamber types.

  10. Principles and procedures of considering item sequence effects in the development of calibrated item pools: Conceptual analysis and empirical illustration

    Directory of Open Access Journals (Sweden)

    Safir Yousfi

    2012-12-01

    Full Text Available Item responses can be context-sensitive. Consequently, composing test forms flexibly from a calibrated item pool requires considering potential context effects. This paper focuses on context effects that are related to the item sequence. It is argued that sequence effects are not necessarily a violation of item response theory but that item response theory offers a powerful tool to analyze them. If sequence effects are substantial, test forms cannot be composed flexibly on the basis of a calibrated item pool, which precludes applications like computerized adaptive testing. In contrast, minor sequence effects do not thwart applications of calibrated item pools. Strategies to minimize the detrimental impact of sequence effects on item parameters are discussed and integrated into a nomenclature that addresses the major features of item calibration designs. An example of an item calibration design demonstrates how this nomenclature can guide the process of developing a calibrated item pool.

  11. Two-Way Satellite Time and Frequency Transfer (TWSTFT) Calibration Constancy From Closure Sums

    Science.gov (United States)

    2008-12-01

    standards in Europe and the US at the 10-15 uncertainty level,” Metrologia , 43, 109-120. [4] D. N. Matsakis, K. Senior, and P. Cook, 2002...2008, “Time transfer with nanosecond accuracy for the realization of International Atomic Time,” Metrologia , 45, 185-198. [7] J. A. Davis et al...The evaluation of uncertainties in [UTC - UTC (k)],” Metrologia , 43, 278-286. 40th Annual Precise Time and Time Interval (PTTI) Meeting 603

  12. Normalization Procedure for the Baptista Depression Scale - Adult Version (EBADEP-A: Transferring of Norms

    Directory of Open Access Journals (Sweden)

    Juliana Oliveira Gomes

    2014-08-01

    Full Text Available Regarding the standardization of psychological instruments, that is, the construction of referential interpretations of a test, we can find different procedures performed both by Classical Test Theory and the Theory of Item Response. Especially in this case (IRT, we can admit a test as a norm, in order to use its standardization and transfer the cut-off point to another instrument. Based on this information, the present study aimed to provide a cutoff score for the Baptista Depression Scale - Adult Version (EBADEP-A through procedures of norms-transfer based on the Center for Epidemiologic Studies – Depression Scale (CES-D. The EBADEP-A presented good distribution and ability to discriminate depressive symptoms, and the sample, consisting of Brazilian College students, received a cutoff score of 32 points. It is emphasized that this is an exploratory and preliminary study, and it we suggest further analyzes to be performed with clinical samples for which results can be corroborated or confronted.

  13. The tabletting machine as an analytical instrument: qualification of the measurement devices for punch forces and validation of the calibration procedures.

    Science.gov (United States)

    Belda, P M; Mielck, J B

    1998-11-01

    The quality of force measurement in an eccentric tabletting machine equipped with piezo-electric load washers mounted under pre-stress at the upper and lower punches, and the reliability of their calibration in situ and under working conditions were carefully investigated, since this tabletting machine is used as an 'analytical instrument' for the evaluation of the compression behaviour of pharmaceutical materials. For a quasistatic calibration procedure the repeatability under standard conditions and the robustness against variations in machine settings, installation conditions, equipment and handling were evaluated. Two differently constructed reference load cells equipped with strain gauges were used for the calibration of the upper punch sensor. The lower punch sensor was calibrated against the upper one. Except for a mechanical hysteresis, owing to uneven stress distribution over the piezo-electric sensors, the results of the quasistatic measurements are assessed to be satisfactory. In addition, dynamic calibrations were performed. One of the strain-gauged load cells was used in addition to two piezo-electric load washers installed without pre-stress. The dynamic behaviour of all the transducers used is deficient. While for the piezo-electric sensors a significant change in the slope of the calibration function with respect to the quasistatic behaviour was observed, for the strain-gauged load cell a pronounced hysteresis must be noted. Comparing the dynamic behaviour at different profiles of rates of force development generated by variations in machine speed and by maximum force setting, the variability in the sensitivity of the upper and lower punch piezo-electric load washers is comparatively small.

  14. Estimating Leaf Bulk Density Distribution in a Tree Canopy Using Terrestrial LiDAR and a Straightforward Calibration Procedure

    Directory of Open Access Journals (Sweden)

    François Pimont

    2015-06-01

    Full Text Available Leaf biomass distribution is a key factor for modeling energy and carbon fluxes in forest canopies and for assessing fire behavior. We propose a new method to estimate 3D leaf bulk density distribution, based on a calibration of indices derived from T-LiDAR. We applied the method to four contrasted plots in a mature Quercus pubescens forest. Leaf bulk densities were measured inside 0.7 m-diameter spheres, referred to as Calibration Volumes. Indices were derived from LiDAR point clouds and calibrated over the Calibration Volume bulk densities. Several indices were proposed and tested to account for noise resulting from mixed pixels and other theoretical considerations. The best index and its calibration parameter were then used to estimate leaf bulk densities at the grid nodes of each plot. These LiDAR-derived bulk density distributions were used to estimate bulk density vertical profiles and loads and above four meters compared well with those assessed by the classical inventory-based approach. Below four meters, the LiDAR-based approach overestimated bulk densities since no distinction was made between wood and leaf returns. The results of our method are promising since they demonstrate the possibility to assess bulk density on small plots at a reasonable operational cost.

  15. Head and neck cancer in elderly patients: is microsurgical free-tissue transfer a safe procedure?

    Science.gov (United States)

    Tarsitano, A; Pizzigallo, A; Sgarzani, R; Oranges, C M; Cipriani, R; Marchetti, C

    2012-12-01

    The safety and success of microvascular transfer have been well documented in the general population, but the good results achieved with the use of free flaps in elderly patients have received little attention. This study sought to identify differences in complications, morbidity and functional outcomes between elderly (≥ 75 years) and younger (patients treated surgically for advanced head and neck cancer using the Head and Neck 35 module of the European Organisation for Research and Treatment of Cancer quality of life questionnaire. Patient treatment consisted of composite resection, including excision of the primary tumour with ipsilateral or bilateral neck dissection and microvascular reconstruction. Eighty-five microvascular tissue transfers were performed to reconstruct major surgical defects. Postoperative radiation therapy was performed when indicated. Total flap loss occurred in three cases in elderly patients and two cases in younger patients. The rates of major surgical complication were 9% in young patients and 11% in elderly patients. No significant difference was observed between the two groups in the rates of major and minor flap complications, morbidity or long-term functional outcome. The results of the present analysis indicate that free-flap microvascular reconstruction can be considered a safe procedure in elderly patients with head and neck cancer.

  16. Transferring the calibration of direct solar irradiance to diffuse-sky radiance measurements for CIMEL Sun-sky radiometers.

    Science.gov (United States)

    Li, Zhengqiang; Blarel, Luc; Podvin, Thierry; Goloub, Philippe; Buis, Jean-Pierre; Morel, Jean-Philippe

    2008-04-01

    Two types of sunphotometric measurement are considered in this study: direct-Sun irradiance and diffuse-sky radiance. Based on CIMEL CE318 Sun-sky radiometer characteristics, we introduce a gain-corrected solid angle that allows interconverting calibration coefficients of these two types of measurement, thus realizing a "vicarious" radiance calibration. The accuracy of the gain-corrected solid angle depends on the number of available historical calibration records. The method is easy to use, provided that at least one laboratory calibration has been made previously. Examples coming from three distinct CE318 versions belonging to the AERONET/PHOTONS network are presented to provide details on the vicarious calibration method and protocols. From the error propagation analysis and the comparison with laboratory results, the uncertainty of the vicarious radiance calibration is shown to be comparable with the laboratory one, e.g., 3%-5%.

  17. SU-E-I-22: Dependence On Calibration Phantom and Field Area of the Conversion Factor Used to Calculate Skin Dose During Neuro-Interventional Fluoroscopic Procedures

    Energy Technology Data Exchange (ETDEWEB)

    Rana, V K; Vijayan, S [Physiology and Biophysics, Toshiba Stroke and Vascular Research Center, University at Buffalo (State University of New York), Buffalo, NY (United States); Rudin, S R; Bednarek, D R [Department of Radiology, Physiology and Biophysics, Toshiba Stroke and Vascular Research Center, University at Buffalo (State University of New York), Buffalo, NY (United States)

    2014-06-01

    Purpose: To determine the appropriate calibration factor to use when calculating skin dose with our real-time dose-tracking system (DTS) during neuro-interventional fluoroscopic procedures by evaluating the difference in backscatter from different phantoms and as a function of entrance-skin field area. Methods: We developed a dose-tracking system to calculate and graphically display the cumulative skin-dose distribution in real time. To calibrate the DTS for neuro-interventional procedures, a phantom is needed that closely approximates the scattering properties of the head. We compared the x-ray backscatter from eight phantoms: 20-cm-thick solid water, 16-cm diameter water-filled container, 16-cm CTDI phantom, modified-ANSI head phantom, 20-cm-thick PMMA, Kyoto-Kagaku PBU- 50 head, Phantom-Labs SK-150 head, and RSD RS-240T head. The phantoms were placed on the patient table with the entrance surface at 15 cm tube-side from the isocenter of a Toshiba Infinix C-arm, and the entrance-skin exposure was measured with a calibrated 6-cc PTW ionization chamber. The measurement included primary radiation, backscatter from the phantom and forward scatter from the table and pad. The variation in entrance-skin exposure was also measured as a function of the skin-entrance area for a 30x30 cm by 20-cm-thick PMMA phantom and the SK-150 head phantom using four different added beam filters. Results: The entranceskin exposure values measured for eight different phantoms differed by up to 12%, while the ratio of entrance exposure of all phantoms relative to solid water showed less than 3% variation with kVp. The change in entrance-skin exposure with entrance-skin area was found to differ for the SK-150 head compared to the 20-cm PMMA phantom and the variation with field area was dependent on the added beam filtration. Conclusion: To accurately calculate skin dose for neuro-interventional procedures with the DTS, the phantom for calibration should be carefully chosen since different

  18. Comparison between Calibration Procedure and Econometric Estimation%校准方法与计量经济方法的比较

    Institute of Scientific and Technical Information of China (English)

    周焯华; 张宗益; 欧阳

    2001-01-01

    The two methods of estimating parameter in computable general equilibrium(CGE) model are introduced and compared:the calibration procedure and econometric estimation. The conclusions are:the estimation of parameter in CGE model must use the calibration procedure coupled with the econometric estimation method;the elasticity of output with respect to labor input,the marginal expenditure share for households and price elasticity of export demand are estimated by econometric estimation method;and other parameters of the CGE model can be get by calibration procedure.%介绍了可计算的一般均衡(CGE)模型中确定参数的两种主要方法:校准方法与计量经济方法,并对两种方法进行了比较,分别论述它们的优缺点。得到的结论是在CGE模型中进行参数估计时采用“校准”方法和计量经济学方法相结合:对于刻画行为人的行为并对结果有重要影响的参数(如各种弹性)采用计量经济学方法来估计,其他的参数则利用所构造的社会核算矩阵(SAM)通过校准方法而得到。

  19. Development of transfer standard devices for ensuring the accurate calibration of ultrasonic physical therapy machines in clinical use

    Energy Technology Data Exchange (ETDEWEB)

    Hekkenberg, R T [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Richards, A [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Beissner, K [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig (Germany); Zeqiri, B [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom); Prout, G [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Cantrall, Ch [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Bezemer, R A [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Koch, Ch [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig, (Germany); Hodnett, M [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom)

    2004-01-01

    Physical therapy ultrasound is widely applied to patients. However, many devices do not comply with the relevant standard stating that the actual power output shall be within {+-}20% of the device indication. Extreme cases have been reported: from delivering effectively no ultrasound or operating at maximum power at all powers indicated. This can potentially lead to patient injury as well as mistreatment. The present European (EC) project is an ongoing attempt to improve the quality of the treatment of patients being treated with ultrasonic physical-therapy. A Portable ultrasound Power Standard (PPS) is being developed and accurately calibrated. The PPS includes: Ultrasound transducers (including one exhibiting an unusual output) and a driver for the ultrasound transducers that has calibration and proficiency test functions. Also included with the PPS is a Cavitation Detector to determine the onset of cavitation occurring within the propagation medium. The PPS will be suitable for conducting in-the-field accreditation (proficiency testing and calibration). In order to be accredited it will be important to be able to show traceability of the calibration, the calibration process and qualification of testing staff. The clinical user will benefit from traceability because treatments will be performed more reliably.

  20. 26 CFR 20.2056A-4T - Procedures for conforming marital trusts and nontrust marital transfers to the requirements of a...

    Science.gov (United States)

    2010-04-01

    ... 26 Internal Revenue 14 2010-04-01 2010-04-01 false Procedures for conforming marital trusts and nontrust marital transfers to the requirements of a qualified domestic trust (temporary). 20.2056A-4T....2056A-4T Procedures for conforming marital trusts and nontrust marital transfers to the requirements...

  1. An Efficient Route to Prepare Metal Organosols by Phase Transfer Procedure

    Institute of Scientific and Technical Information of China (English)

    Ji Kuan ZHAO; Xiao CHEN; Chun Jie YANG; Zhen Ming SUI; Yong Cun CHAI; Guo Dong ZHANG; Jie LIU

    2005-01-01

    Silver and gold organosols are easily prepared by transferring nanoparticles from aqueous phase into isooctane with high efficiency (>90%). Concentrations of sodium oleate and magnesium chloride have crucial effects on the transfer efficiency. Based on the UV-visible absorption spectra, TEM micrographs of nanoparticles, as well as molecular modeling calculation about the adsorption conformation of sodium oleate molecules, a possible phase transfer mechanism is proposed.

  2. Two-demensional analysis of heat and mass transfer in porous media using the strongly implicit procedure

    Science.gov (United States)

    Curry, D. M.

    1974-01-01

    Numerical results of the heat and mass transfer in a porous matrix are presented. The coupled, nonlinear partial differential equations describing this physical phenomenon are solved in finite difference form for two dimensions, using a new iterative technique (the strongly implicit procedure). The influence of the external environment conditions (heating and pressure) is shown to produce two-dimensional flow in the porous matrix. Typical fluid and solid temperature distributions in the porous matrix and internal pressure distributions are presented.

  3. The Landsat Data Continuity Mission Operational Land Imager (OLI) Radiometric Calibration

    Science.gov (United States)

    Markham, Brian L.; Dabney, Philip W.; Murphy-Morris, Jeanine E.; Knight, Edward J.; Kvaran, Geir; Barsi, Julia A.

    2010-01-01

    The Operational Land Imager (OLI) on the Landsat Data Continuity Mission (LDCM) has a comprehensive radiometric characterization and calibration program beginning with the instrument design, and extending through integration and test, on-orbit operations and science data processing. Key instrument design features for radiometric calibration include dual solar diffusers and multi-lamped on-board calibrators. The radiometric calibration transfer procedure from NIST standards has multiple checks on the radiometric scale throughout the process and uses a heliostat as part of the transfer to orbit of the radiometric calibration. On-orbit lunar imaging will be used to track the instruments stability and side slither maneuvers will be used in addition to the solar diffuser to flat field across the thousands of detectors per band. A Calibration Validation Team is continuously involved in the process from design to operations. This team uses an Image Assessment System (IAS), part of the ground system to characterize and calibrate the on-orbit data.

  4. Device-Task Fidelity and Transfer of Training: Aircraft Cockpit Procedures Training.

    Science.gov (United States)

    Prophet, Wallace W.; Boyd, H. Alton

    An evaluation was made of the training effectiveness of two cockpit procedures training devices, differing greatly in physical fidelity and cost, for use on the ground for a twin-engine, turboprop, fixed-wing aircraft. One group of students received training in cockpit procedures in a relatively expensive, sophisticated, computerized trainer,…

  5. Evaluation of procedural learning transfer from a virtual environment to a real situation: a case study on tank maintenance training.

    Science.gov (United States)

    Ganier, Franck; Hoareau, Charlotte; Tisseau, Jacques

    2014-01-01

    Virtual reality opens new opportunities for operator training in complex tasks. It lowers costs and has fewer constraints than traditional training. The ultimate goal of virtual training is to transfer knowledge gained in a virtual environment to an actual real-world setting. This study tested whether a maintenance procedure could be learnt equally well by virtual-environment and conventional training. Forty-two adults were divided into three equally sized groups: virtual training (GVT® [generic virtual training]), conventional training (using a real tank suspension and preparation station) and control (no training). Participants then performed the procedure individually in the real environment. Both training types (conventional and virtual) produced similar levels of performance when the procedure was carried out in real conditions. Performance level for the two trained groups was better in terms of success and time taken to complete the task, time spent consulting job instructions and number of times the instructor provided guidance.

  6. SeaWiFS Postlaunch Technical Report Series. Volume 5; The SeaWiFS Solar Radiation-Based Calibration and the Transfer-to-Orbit Experiment

    Science.gov (United States)

    Hooker, Stanford B. (Editor); Firestone, Elaine R. (Editor); Barnes, Robert A.; Eplee, Robert E., Jr.; Biggar, Stuart F.; Thome, Kurtis J.; Zalewski, Edward F.; Slater, Philip N.; Holmes, Alan W.

    1999-01-01

    The solar radiation-based calibration (SRBC) of the Sea-viewing Wide Field-of-view Sensor (SeaWiFS) was performed on 1 November 1993. Measurements were made outdoors in the courtyard of the instrument manufacturer. SeaWiFS viewed the solar irradiance reflected from the sensor's diffuser in the same manner as viewed on orbit. The calibration included measurements using a solar radiometer designed to determine the transmittances of principal atmospheric constituents. The primary uncertainties in the outdoor measurements are the transmission of the atmosphere and the reflectance of the diffuser. Their combined uncertainty is about 5 or 6%. The SRBC also requires knowledge of the extraterrestrial solar spectrum. Four solar models are used. When averaged over the responses of the SeaWiFS bands, the irradiance models agree at the 3.6% level, with the greatest difference for SeaWiFS band 8. The calibration coefficients from the SRBC are lower than those from the laboratory calibration of the instrument in 1997. For a representative solar model, the ratios of the SRBC coefficients to laboratory values average 0.962 with a standard deviation of 0.012. The greatest relative difference is 0.946 for band 8. These values are within the estimated uncertainties of the calibration measurements. For the transfer-to-orbit experiment, the measurements in the manufacturer's courtyard are used to predict the digital counts from the instrument on its first day on orbit (August 1, 1997). This experiment requires an estimate of the relative change in the diffuser response for the period between the launch of the instrument and its first solar measurements on orbit (September 9, 1997). In relative terms, the counts from the instrument on its first day on orbit averaged 1.3% higher than predicted, with a standard deviation of 1.2% and a greatest difference of 2.4% or band 7. The estimated uncertainty for the transfer-to-orbit experiment is about 3 or 4%.

  7. Providing radiometric traceability for the calibration home base of DLR by PTB

    Energy Technology Data Exchange (ETDEWEB)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Hoepe, A.; Hauer, K.-O. [Physikalisch-Technische Bundesanstalt, Braunschweig und Berlin, 10587 Berlin (Germany); Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A. [Deutsches Zentrum fuer Luft- und Raumfahrt, Institut fuer Methodik der Fernerkundung, 82234 Oberpfaffenhofen (Germany)

    2013-05-10

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance E{sub {lambda}}({lambda}) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor {beta}{sub 0 Degree-Sign :45 Degree-Sign }({lambda}) of the RASTA reflection standard was calibrated in a 0 Degree-Sign :45 Degree-Sign -viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k= 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0 Degree-Sign :45 Degree-Sign -viewing geometry spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign }({lambda}) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k= 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign ,calc}({lambda}) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k= 2) of both procedures.

  8. Providing radiometric traceability for the calibration home base of DLR by PTB

    Science.gov (United States)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Höpe, A.; Hauer, K.-O.; Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A.

    2013-05-01

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance Eλ(λ) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor β0°:45°(λ) of the RASTA reflection standard was calibrated in a 0°:45°-viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k = 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0°:45°-viewing geometry spectral radiance Lλ,0°:45°(λ) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k = 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance Lλ,0°:45°,calc(λ) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k = 2) of both procedures.

  9. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Cortes P, A.; Garcia D, O.C.; Ortiz P, I.; Becerril V, A. [Instituto nacional de Investigaciones Nucleares, Laboratorio de Patrones Radiactivos, Departamento de Metrologia, A.P. 18-1027. C.P. 11801 Mexico D.F. (Mexico)

    2000-07-01

    With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup 3} contained in polyethylene small bottles of 25 cm{sup 3} elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm{sup 3} of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)

  10. A Probe Trial Procedure for the Study of Children's Discrimination Learning and Transfer

    Science.gov (United States)

    Cole, Michael

    1976-01-01

    A discrimination reversal problem was presented to 192 children varying in age from 3 to 5 years. At the end of both the initial learning and transfer trials, probe trials were introduced to ascertain the response rule describing children's choices. (Author/SB)

  11. OLI Radiometric Calibration

    Science.gov (United States)

    Markham, Brian; Morfitt, Ron; Kvaran, Geir; Biggar, Stuart; Leisso, Nathan; Czapla-Myers, Jeff

    2011-01-01

    Goals: (1) Present an overview of the pre-launch radiance, reflectance & uniformity calibration of the Operational Land Imager (OLI) (1a) Transfer to orbit/heliostat (1b) Linearity (2) Discuss on-orbit plans for radiance, reflectance and uniformity calibration of the OLI

  12. Dry calibration of ultrasonic gas flow meters

    Energy Technology Data Exchange (ETDEWEB)

    De Boer, G.; Lansing, J.

    1997-07-01

    At present in most European countries it is customary that turbine meters, or the newer ultrasonic gas flow meters, when used in fiscal metering or custody transfer metering applications, are calibrated in a test facility by comparison to standards or reference devices. For reason of practical and operational drawbacks, costs involved and availability of only a limited number of calibration facilities, another way of meter verification is advantageous. For orifice metering the practice of dry calibration is well established; that is, meter verification is based upon examination of the geometry and installation of the orifice plate and a function check of the read out devices. Although for turbine meters a flow (wet) calibration may be a necessity, it will be shown that ultrasonic gas flow meters can be dry calibrated in the same way as orifice meters. As a basis for the acceptance of a dry calibration procedure for ultrasonic gas flow meters, a sensitivity analysis of the relevant variables with respect to the meter's accuracy is presented. Further test results are presented that demonstrate the feasibility of the concept of dry calibration applied to ultrasonic gas flow meters. (author)

  13. Do Basic Psychomotor Skills Transfer Between Different Image-based Procedures?

    NARCIS (Netherlands)

    Buzink, S.N.; Goossens, R.H.M.; Schoon, E.J.; De Ridder, H.; Jakimowicz, J.J.

    2010-01-01

    Background - Surgical techniques that draw from multiple types of image-based procedures (IBP) are increasing, such as Natural Orifice Transluminal Endoscopic Surgery, fusing laparoscopy and flexible endoscopy. However, little is known about the relation between psychomotor skills for performing dif

  14. Calibrating the Station Biases for the C-Band Transfer Measuring System%C波段转发测轨系统测站偏差的标校

    Institute of Scientific and Technical Information of China (English)

    宋小勇; 毛悦; 贾小林

    2012-01-01

    C波段转发测轨体制在GE~卫星测定轨中有突出优势,但其测量系统偏差的标校精度对定轨结果影响显著。基于L波段载波相位多星定轨结果,提出一种基于标准轨道拟合残差的系统偏差标校方法。该方法能够同时标校测量偏差及时标偏差,有利于弥补激光测距标校方法观测量较少的缺陷。利用国内监测站对中国导航卫星实测数据分析结果表明,经过标校后的C波段转发测轨精度可优于5m,测量偏差及时标偏差具有较好的稳定性。%The C-band transfer measuring system plays an important role in the precise orbit determination (POD) for geostationary satellite. However, the performance of the system was remarkably influenced by the accuracies of the station biases. Based on the orbits from the multi-satellite POD using the carrier observation of L-band, a new method to calibrate the biases from orbit fitting of C-band measurement is presented. This method can not only estimate the ranging biases and time biases simultaneously, but also overcome the shortcoming of too few observa- tions when using the SLR calibrating method. The POD experiment of Compass navigation satellite shows that the satellite~s position error is about 5 m after the station biases is calibrated, and the estimated ranging biases and time biases have higher stability and reliability.

  15. Diabetic limb salvage procedure with bone allograft and free flap transfer: a case report

    Science.gov (United States)

    Godoy-Santos, Alexandre L.; Amodio, Daniel T.; Pires, André; Lima, Ana L. M.; Wei, Teng H.; de Cesar-Netto, Cesar; Armstrong, David G.

    2017-01-01

    ABSTRACT The aim of this case report was to describe a successful diabetic limb salvage procedure in the treatment of an infected diabetic foot ulcer through a multidisciplinary team approach and complex surgical reconstruction involving a femoral head bone allograft and musculocutaneous latissimus dorsi free flap. The decision to proceed with aggressive staged efforts at diabetic limb salvage should be made only after careful consultation with the patient, his or her family, and the rest of the multidisciplinary healthcare team.

  16. Calibration and temperature profile of a tungsten filament lamp

    Energy Technology Data Exchange (ETDEWEB)

    De Izarra, Charles [Groupe de Recherche sur l' Energetique des Milieux Ionises, UMR6606 Universite d' Orleans, CNRS, Faculte des Sciences, Site de Bourges, rue Gaston Berger, BP 4043, 18028 Bourges Cedex (France); Gitton, Jean-Michel, E-mail: Charles.De_Izarra@univ-orleans.f [College Littre, 10 rue Littre, Bourges (France)

    2010-07-15

    The goal of this work proposed for undergraduate students and teachers is the calibration of a tungsten filament lamp from electric measurements that are both simple and precise, allowing to determine the temperature of tungsten filament as a function of the current intensity. This calibration procedure was first applied to a conventional filament lamp (lamp used in automotive lighting) and then tested on a standard tungsten ribbon lamp. The calibration procedure developed was checked by determining the calibration point of the tungsten ribbon lamp with an accuracy of 2%. In addition, for low current intensity, it was observed that the temperature of the filament was not uniform; an explanation is proposed by considering a simple heat transfer model.

  17. 26 CFR 20.2056A-4 - Procedures for conforming marital trusts and nontrust marital transfers to the requirements of a...

    Science.gov (United States)

    2010-04-01

    ... 26 Internal Revenue 14 2010-04-01 2010-04-01 false Procedures for conforming marital trusts and nontrust marital transfers to the requirements of a qualified domestic trust. 20.2056A-4 Section 20.2056A-4... conforming marital trusts and nontrust marital transfers to the requirements of a qualified domestic...

  18. In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals

    Directory of Open Access Journals (Sweden)

    Webster John G

    2006-10-01

    Full Text Available Abstract Background We need a sensor to measure the convective heat transfer coefficient during ablation of the heart or liver. Methods We built a minimally invasive instrument to measure the in vivo convective heat transfer coefficient, h in animals, using a Wheatstone-bridge circuit, similar to a hot-wire anemometer circuit. One arm is connected to a steerable catheter sensor whose tip is a 1.9 mm × 3.2 mm thin film resistive temperature detector (RTD sensor. We used a circulation system to simulate different flow rates at 39°C for in vitro experiments using distilled water, tap water and saline. We heated the sensor approximately 5°C above the fluid temperature. We measured the power consumed by the sensor and the resistance of the sensor during the experiments and analyzed these data to determine the value of the convective heat transfer coefficient at various flow rates. Results From 0 to 5 L/min, experimental values of h in W/(m2·K were for distilled water 5100 to 13000, for tap water 5500 to 12300, and for saline 5400 to 13600. Theoretical values were 1900 to 10700. Conclusion We believe this system is the smallest, most accurate method of minimally invasive measurement of in vivo h in animals and provides the least disturbance of flow.

  19. The Science of Calibration

    Science.gov (United States)

    Kent, S. M.

    2016-05-01

    This paper presents a broad overview of the many issues involved in calibrating astronomical data, covering the full electromagnetic spectrum from radio waves to gamma rays, and considering both ground-based and space-based missions. These issues include the science drivers for absolute and relative calibration, the physics behind calibration and the mechanisms used to transfer it from the laboratory to an astronomical source, the need for networks of calibrated astronomical standards, and some of the challenges faced by large surveys and missions.

  20. Modeling Agricultural Watersheds with the Soil and Water Assessment Tool (SWAT): Calibration and Validation with a Novel Procedure for Spatially Explicit HRUs.

    Science.gov (United States)

    Teshager, Awoke Dagnew; Gassman, Philip W; Secchi, Silvia; Schoof, Justin T; Misgna, Girmaye

    2016-04-01

    Applications of the Soil and Water Assessment Tool (SWAT) model typically involve delineation of a watershed into subwatersheds/subbasins that are then further subdivided into hydrologic response units (HRUs) which are homogeneous areas of aggregated soil, landuse, and slope and are the smallest modeling units used within the model. In a given standard SWAT application, multiple potential HRUs (farm fields) in a subbasin are usually aggregated into a single HRU feature. In other words, the standard version of the model combines multiple potential HRUs (farm fields) with the same landuse/landcover, soil, and slope, but located at different places of a subbasin (spatially non-unique), and considers them as one HRU. In this study, ArcGIS pre-processing procedures were developed to spatially define a one-to-one match between farm fields and HRUs (spatially unique HRUs) within a subbasin prior to SWAT simulations to facilitate input processing, input/output mapping, and further analysis at the individual farm field level. Model input data such as landuse/landcover (LULC), soil, crop rotation, and other management data were processed through these HRUs. The SWAT model was then calibrated/validated for Raccoon River watershed in Iowa for 2002-2010 and Big Creek River watershed in Illinois for 2000-2003. SWAT was able to replicate annual, monthly, and daily streamflow, as well as sediment, nitrate and mineral phosphorous within recommended accuracy in most cases. The one-to-one match between farm fields and HRUs created and used in this study is a first step in performing LULC change, climate change impact, and other analyses in a more spatially explicit manner.

  1. Modeling Agricultural Watersheds with the Soil and Water Assessment Tool (SWAT): Calibration and Validation with a Novel Procedure for Spatially Explicit HRUs

    Science.gov (United States)

    Teshager, Awoke Dagnew; Gassman, Philip W.; Secchi, Silvia; Schoof, Justin T.; Misgna, Girmaye

    2016-04-01

    Applications of the Soil and Water Assessment Tool (SWAT) model typically involve delineation of a watershed into subwatersheds/subbasins that are then further subdivided into hydrologic response units (HRUs) which are homogeneous areas of aggregated soil, landuse, and slope and are the smallest modeling units used within the model. In a given standard SWAT application, multiple potential HRUs (farm fields) in a subbasin are usually aggregated into a single HRU feature. In other words, the standard version of the model combines multiple potential HRUs (farm fields) with the same landuse/landcover, soil, and slope, but located at different places of a subbasin (spatially non-unique), and considers them as one HRU. In this study, ArcGIS pre-processing procedures were developed to spatially define a one-to-one match between farm fields and HRUs (spatially unique HRUs) within a subbasin prior to SWAT simulations to facilitate input processing, input/output mapping, and further analysis at the individual farm field level. Model input data such as landuse/landcover (LULC), soil, crop rotation, and other management data were processed through these HRUs. The SWAT model was then calibrated/validated for Raccoon River watershed in Iowa for 2002-2010 and Big Creek River watershed in Illinois for 2000-2003. SWAT was able to replicate annual, monthly, and daily streamflow, as well as sediment, nitrate and mineral phosphorous within recommended accuracy in most cases. The one-to-one match between farm fields and HRUs created and used in this study is a first step in performing LULC change, climate change impact, and other analyses in a more spatially explicit manner.

  2. Calibrating photon counts from a single image

    CERN Document Server

    Heintzmann, Rainer; Nieuwenhuizen, Robert P J; Lidke, Keith A; Rieger, Bernd

    2016-01-01

    Most image capturing devices do not directly report the number of detected photons, but a value proportional to the photoelectron charge produced in a photomultiplier tube or collected in a camera pixel. In order to establish the photon count, the gain of the device must be measured, typically by recording tens of calibration images and exploiting the linear relationship between mean intensity and its variance [vanVliet1998]. Here we propose and evaluate a method that obtains the gain from a single acquired image by quantifying out-of-band information. As noise is not limited to the cut-off frequency of the optical transfer function (OTF), estimation of the out-of-band energy relative to the total energy enables computation of the gain. We show on simulation and experimental data that this much simpler procedure, which can be retroactively applied to any image, is comparable in precision to traditional gain calibration procedures.

  3. Calibration of Galileo signals for time metrology.

    Science.gov (United States)

    Defraigne, Pascale; Aerts, Wim; Cerretto, Giancarlo; Cantoni, Elena; Sleewaegen, Jean-Marie

    2014-12-01

    Using global navigation satellite system (GNSS) signals for accurate timing and time transfer requires the knowledge of all electric delays of the signals inside the receiving system. GNSS stations dedicated to timing or time transfer are classically calibrated only for Global Positioning System (GPS) signals. This paper proposes a procedure to determine the hardware delays of a GNSS receiving station for Galileo signals, once the delays of the GPS signals are known. This approach makes use of the broadcast satellite inter-signal biases, and is based on the ionospheric delay measured from dual-frequency combinations of GPS and Galileo signals. The uncertainty on the so-determined hardware delays is estimated to 3.7 ns for each isolated code in the L5 frequency band, and 4.2 ns for the ionosphere-free combination of E1 with a code of the L5 frequency band. For the calibration of a time transfer link between two stations, another approach can be used, based on the difference between the common-view time transfer results obtained with calibrated GPS data and with uncalibrated Galileo data. It is shown that the results obtained with this approach or with the ionospheric method are equivalent.

  4. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    Energy Technology Data Exchange (ETDEWEB)

    DeWard, L.A.; Micka, J.A. [Univ. of Wisconsin, Madison, WI (United States)

    1993-12-31

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST).

  5. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  6. Calibrating nacelle lidars

    OpenAIRE

    Courtney, Michael

    2013-01-01

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail.The first of these is a line of sight...

  7. Transfer of Salmonella enterica Serovar Typhimurium from Beef to Tomato through Kitchen Equipment and the Efficacy of Intermediate Decontamination Procedures.

    Science.gov (United States)

    Gkana, E; Lianou, A; Nychas, G-J E

    2016-07-01

    It is well established that a high percentage of foodborne illness is caused by failure of consumers to prepare food in a hygienic manner. Indeed, a common practice in households is to use the same kitchen equipment for both raw meat and fresh produce. Such a practice may lead to cross-contamination of fruits and vegetables, which are mainly consumed without further processing, with pathogenic microorganisms originating from raw meat. The present study was performed to examine the transfer of the pathogenic bacterium Salmonella enterica serovar Typhimurium from inoculated beef fillets to tomatoes via contact with high-density polyethylene (PE), stainless steel (SS), and wooden (WD) surfaces and through cutting with SS knives. Furthermore, the following decontamination procedures were applied: (i) rinsing with tap water, (ii) scrubbing with tap water and liquid dish detergent, and (iii) using a commercial antibacterial spray. When surfaces and knives that came into contact with contaminated beef fillets were not cleaned prior to handling tomatoes, the lowest level of pathogen transfer to tomatoes was observed through PE surfaces. All of the decontamination procedures applied were more effective on knives than on surfaces, while among the surface materials tested, WD surfaces were the most difficult to decontaminate, followed by PE and SS surfaces. Mechanical cleaning with tap water and detergent was more efficient in decontaminating WD surfaces than using commercial disinfectant spray, followed by rinsing only with water. Specifically, reductions of 2.07 and 1.09 log CFU/cm(2) were achieved by washing the WD surfaces with water and detergent and spraying the surfaces with an antibacterial product, respectively. Although the pathogen's populations on SS and PE surfaces, as well as on tomatoes, after both aforementioned treatments were under the detection limit, the surfaces were all positive after enrichment, and thus, the potential risk of cross-contamination cannot

  8. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  9. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration unce...

  10. Calibration report for Avent 5-beam Demonstrator lidar

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael

    . Their estimation is the ultimate goal of a calibration: a relation is established between reference measurements from calibrated instruments and corresponding LiDAR indications. Traceability in the calibration is obtained by transferring measurement uncertainties from the reference instrument through...

  11. A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator.

    Science.gov (United States)

    Ceccatelli, A; Campanella, F; Ciofetta, G; Marracino, F M; Cannatà, V

    2010-02-01

    To determine photopeak efficiency for (99m)Tc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected.

  12. Calibrating nacelle lidars

    DEFF Research Database (Denmark)

    Courtney, Michael

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report...... presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated...... a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam...

  13. Internet-based calibration of a multifunction calibrator

    Energy Technology Data Exchange (ETDEWEB)

    BUNTING BACA,LISA A.; DUDA JR.,LEONARD E.; WALKER,RUSSELL M.; OLDHAM,NILE; PARKER,MARK

    2000-04-17

    A new way of providing calibration services is evolving which employs the Internet to expand present capabilities and make the calibration process more interactive. Sandia National Laboratories and the National Institute of Standards and Technology are collaborating to set up and demonstrate a remote calibration of multifunction calibrators using this Internet-based technique that is becoming known as e-calibration. This paper describes the measurement philosophy and the Internet resources that can provide real-time audio/video/data exchange, consultation and training, as well as web-accessible test procedures, software and calibration reports. The communication system utilizes commercial hardware and software that should be easy to integrate into most calibration laboratories.

  14. Calibrating transition-metal energy levels and oxygen bands in first-principles calculations: Accurate prediction of redox potentials and charge transfer in lithium transition-metal oxides

    Science.gov (United States)

    Seo, Dong-Hwa; Urban, Alexander; Ceder, Gerbrand

    2015-09-01

    Transition-metal (TM) oxides play an increasingly important role in technology today, including applications such as catalysis, solar energy harvesting, and energy storage. In many of these applications, the details of their electronic structure near the Fermi level are critically important for their properties. We propose a first-principles-based computational methodology for the accurate prediction of oxygen charge transfer in TM oxides and lithium TM (Li-TM) oxides. To obtain accurate electronic structures, the Heyd-Scuseria-Ernzerhof (HSE06) hybrid functional is adopted, and the amount of exact Hartree-Fock exchange (mixing parameter) is adjusted to reproduce reference band gaps. We show that the HSE06 functional with optimal mixing parameter yields not only improved electronic densities of states, but also better energetics (Li-intercalation voltages) for LiCo O2 and LiNi O2 as compared to the generalized gradient approximation (GGA), Hubbard U corrected GGA (GGA +U ), and standard HSE06. We find that the optimal mixing parameters for TM oxides are system specific and correlate with the covalency (ionicity) of the TM species. The strong covalent (ionic) nature of TM-O bonding leads to lower (higher) optimal mixing parameters. We find that optimized HSE06 functionals predict stronger hybridization of the Co 3 d and O 2 p orbitals as compared to GGA, resulting in a greater contribution from oxygen states to charge compensation upon delithiation in LiCo O2 . We also find that the band gaps of Li-TM oxides increase linearly with the mixing parameter, enabling the straightforward determination of optimal mixing parameters based on GGA (α =0.0 ) and HSE06 (α =0.25 ) calculations. Our results also show that G0W0@GGA +U band gaps of TM oxides (M O ,M =Mn ,Co ,Ni ) and LiCo O2 agree well with experimental references, suggesting that G0W0 calculations can be used as a reference for the calibration of the mixing parameter in cases when no experimental band gap has been

  15. Procedures for calibration of clinical dosemeters of radiotherapy with gamma radiation Co-60 in the IPEN-CNEN, Sao Paulo, Brazil;Procedimento de calibracao de dosimetros clinicos de radioterapia em feixes de radiacao gamma de Co-60 no IPEN-CNEN/SP

    Energy Technology Data Exchange (ETDEWEB)

    Damatto, Willian B.; Santos, Gelson P.; Potiens, Maria da Penha A.; Vivolo, Vitor [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2009-07-01

    The periodic calibration of clinic dosimeters is a very important task, because those measurement systems performing the control quality of the radioactive source and for that assure the control quality of the patient dose estimative. This paper shows the calibration procedures and routine improved in the calibration laboratory of IPEN/CNEN-SP to be applied to the clinic dosimeters following international and national recommendations and yet the spend care that the clients should have with the utilization of that measurement systems for control quality of the radiotherapy procedures, for reduce the dose received for patient, and workers, and for increase the confidence level of the measurements.(author)

  16. Protein standardization IV: Value transfer procedure for the assignment of serum protein values from a reference preparation to a target material.

    Science.gov (United States)

    Blirup-Jensen, S; Johnson, A M; Larsen, M

    2001-11-01

    A new approach for the assignment of values to serum proteins in a target material using a reference preparation has been developed. The procedure describes the general as well as the practical principles involved in the value assignment (with examples). Two models have been developed: 1) The direct value transfer between serum matrices and 2) the indirect value transfer from a pure protein preparation to a serum protein material. The necessary mathematical equations are developed and explained. The data reduction and statistical evaluation are discussed. The practical procedure (the transfer protocol) is based on six dilutions of the reference preparation assayed together with six dilutions of the target material. In this way imprecision is reduced and the proportionality of the two materials (i.e. the presence or absence of matrix effects) can be assessed directly by evaluating a single regression plot. If no matrix effects are found, the regression line will pass through zero with a slope equal to the ratio of the concentrations of the two materials. The transfer protocol is based on a multiple point value assignment obtained by several measurements a day repeated on several days, an important prerequisite being that all reconstitutions and dilutions are controlled by weighing.

  17. "Calibration-on-the-spot": How to calibrate an EMCCD camera from its images.

    Science.gov (United States)

    Mortensen, Kim I; Flyvbjerg, Henrik

    2016-07-06

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information.

  18. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  19. Calibration of environmental radionuclide transfer models using a Bayesian approach with Markov chain Monte Carlo simulations and model comparisons - Calibration of radionuclides transfer models in the environment using a Bayesian approach with Markov chain Monte Carlo simulation and comparison of models

    Energy Technology Data Exchange (ETDEWEB)

    Nicoulaud-Gouin, V.; Giacalone, M.; Gonze, M.A. [Institut de Radioprotection et de Surete Nucleaire-PRP-ENV/SERIS/LM2E (France); Martin-Garin, A.; Garcia-Sanchez, L. [IRSN-PRP-ENV/SERIS/L2BT (France)

    2014-07-01

    Calibration of transfer models according to observation data is a challenge, especially if parameters uncertainty is required, and if competing models should be decided between them. Generally two main calibration methods are used: The frequentist approach in which the unknown parameter of interest is supposed fixed and its estimation is based on the data only. In this category, least squared method has many restrictions in nonlinear models and competing models need to be nested in order to be compared. The bayesian inference in which the unknown parameter of interest is supposed random and its estimation is based on the data and on prior information. Compared to frequentist method, it provides probability density functions and therefore pointwise estimation with credible intervals. However, in practical cases, Bayesian inference is a complex problem of numerical integration, which explains its low use in operational modeling including radioecology. This study aims to illustrate the interest and feasibility of Bayesian approach in radioecology particularly in the case of ordinary differential equations with non-constant coefficients models, which cover most radiological risk assessment models, notably those implemented in the Symbiose platform (Gonze et al, 2010). Markov Chain Monte Carlo (MCMC) method (Metropolis et al., 1953) was used because the posterior expectations are intractable integrals. The invariant distribution of the parameters was performed by the metropolis-Hasting algorithm (Hastings, 1970). GNU-MCSim software (Bois and Maszle, 2011) a bayesian hierarchical framework, was used to deal with nonlinear differential models. Two case studies including this type of model were investigated: An Equilibrium Kinetic sorption model (EK) (e.g. van Genuchten et al, 1974), with experimental data concerning {sup 137}Cs and {sup 85}Sr sorption and desorption in different soils studied in stirred flow-through reactors. This model, generalizing the K{sub d} approach

  20. Intercomparison of calibration procedures of high dose rate {sup 192} Ir sources in Brazil and a proposal of a new methodology; Intercomparacao de procedimientos de calibracao de fontes de {sup 192} Ir de alta taxa de dose no Brasil e proposta de uma nova metodologia

    Energy Technology Data Exchange (ETDEWEB)

    Marechal, M.H.; Almeida, C.E. de [Laboratorio Nacional de Metrologia das Radiacoes Ionizantes IRD/CNEN. Caixa Postal 37750 CEP 22780-160 Rio de Janeiro (Brazil)

    1998-12-31

    The objective of this paper is to report the results of an intercomparison of the calibration procedures for {sup 192} Ir sources presently in use in Brazil and to proposal a calibration procedure to derive the N{sub k} for a Farmer type ionization chamber for {sup 192} Ir energy by interpolating from a {sup 60} Co gamma-rays and 250 kV x-rays calibration factors. the intercomparison results were all within {+-} 3.0 % except one case where 4.6 % was observed and latter identified as a problem with N-k value for X-rays. The method proposed by the present work make possible the improvement of the metrological coherence among the calibration laboratories and their users once the N{sub k} values could then provided by any of the members of SSDL network. (Author)

  1. Versatile procedure for site-specific grafting of polymer brushes on patchy particles via atom transfer radical polymerization (ATRP)

    NARCIS (Netherlands)

    Van Ravensteijn, Bas G P; Kegel, Willem K.

    2016-01-01

    We report the preparation of chemically anisotropic colloidal dumbbells of which one lobe is functionalized with chemical handles in the form of chlorine groups. The chlorines are easily converted to azides and subsequently to active initiators for Atom Transfer Radical Polymerization (ATRP) by Clic

  2. THE METHOD OF DETERMINATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. Part 3. The dependence of the radiation beam characteristics

    Directory of Open Access Journals (Sweden)

    E. V. Titovich

    2016-01-01

    Full Text Available To ensure the radiation protection of oncology patients is needed to provide the constancy of functional characteristics of the medical linear accelerators, which affect the accuracy of dose delivery. For this purpose, their quality control procedures are realized including calibration of radiation output of the linac, so the error in determining the dose reference value during this procedure must not exceed 2 %. The aim is to develop a methodology for determining the error in determining this value, depending on the characteristics of the radiation beam. Dosimetric measurements of Trilogy S/N 3567 linac dose distributions have been carried out for achievement of the objectives, on the basis of which dose errors depending on the dose rate value, the accuracy of the beam quality and output factors determination, the symmetry and uniformity of the radiation field, the angular dependence of the linac radiation output were obtained. It was found that the greatest impact on the value of the error has the error in the output factors determination (up to 5.26 % for both photon energy. Dose errors caused by changing dose rate during treatment were different for two photon energies, and reached 1.6 % for 6 MeV and 1.4 % for 18 MeV. Dose errors caused by inaccuracies of the beam quality determination were different for two photon energies, and reached 1.1 % for 18 MeV and –0.3 % for 6 MeV. Errors caused by the remaining of the characteristic do not exceed 1 %. Thus, there is a possibility to express the results of periodic quality control of the linear accelerator in terms of dose and use them to conduct a comprehensive assessment of the possibility of clinical use of a linear accelerator for oncology patients irradiation on the basis of the calibration of radiation output.

  3. Calculation procedure to determine average mass transfer coefficients in packed columns from experimental data for ammonia-water absorption refrigeration systems

    Energy Technology Data Exchange (ETDEWEB)

    Sieres, Jaime; Fernandez-Seara, Jose [University of Vigo, Area de Maquinas y Motores Termicos, E.T.S. de Ingenieros Industriales, Vigo (Spain)

    2008-08-15

    The ammonia purification process is critical in ammonia-water absorption refrigeration systems. In this paper, a detailed and a simplified analytical model are presented to characterize the performance of the ammonia rectification process in packed columns. The detailed model is based on mass and energy balances and simultaneous heat and mass transfer equations. The simplified model is derived and compared with the detailed model. The range of applicability of the simplified model is determined. A calculation procedure based on the simplified model is developed to determine the volumetric mass transfer coefficients in the vapour phase from experimental data. Finally, the proposed model and other simple calculation methods found in the general literature are compared. (orig.)

  4. Lidar to lidar calibration phase 2

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the results from phase 2 of a lidar to lidar (L2L) calibration procedure. Phase two of the project included two measurement campaigns conducted at given sites. The purpose was to find out if the lidar-to-lidar calibration procedure can be conducted with similar results...

  5. System for measuring the effect of fouling and corrosion on heat transfer under simulated OTEC conditions. [HTAU and LABTTF codes

    Energy Technology Data Exchange (ETDEWEB)

    Fetkovich, J.G.

    1976-12-01

    A complete system designed to measure, with high precision, changes in heat transfer rates due to fouling and corrosion of simulated heat exchanger tubes, at sea and under OTEC conditions is described. All aspects of the system are described in detail, including theory, mechanical design, electronics design, assembly procedures, test and calibration, operating procedures, laboratory results, field results, and data analysis programs.

  6. Optical Calibration For Jefferson Lab HKS Spectrometer

    Energy Technology Data Exchange (ETDEWEB)

    L. Yuan; L. Tang

    2005-11-04

    In order to accept very forward angle scattering particles, Jefferson Lab HKS experiment uses an on-target zero degree dipole magnet. The usual spectrometer optics calibration procedure has to be modified due to this on-target field. This paper describes a new method to calibrate HKS spectrometer system. The simulation of the calibration procedure shows the required resolution can be achieved from initially inaccurate optical description.

  7. Observatory Magnetometer In-Situ Calibration

    Directory of Open Access Journals (Sweden)

    A Marusenkov

    2011-07-01

    Full Text Available An experimental validation of the in-situ calibration procedure, which allows estimating parameters of observatory magnetometers (scale factors, sensor misalignment without its operation interruption, is presented. In order to control the validity of the procedure, the records provided by two magnetometers calibrated independently in a coil system have been processed. The in-situ estimations of the parameters are in very good agreement with the values provided by the coil system calibration.

  8. Proximal ulnar stump stability after using the pronator quadratus muscle transfer combined with the Suavé-Kapandji procedure in rheumatoid wrist.

    Science.gov (United States)

    Uerpairojkit, Chairoj; Leechavengvongs, Somsak; Malungpaishorpe, Kanchai; Witoonchart, Kiat; Buddhavibul, Panai

    2014-01-01

    The pronator quadratus muscle transfer combined with the Sauvé-Kapandji procedure was used to treat the distal radioulnar joint disorder in ten rheumatoid wrists for prevention against instability of the proximal ulnar stump. All patients were female with a mean age of 46.6 years. The mean follow-up time was 24.2 months. Postoperatively, supination increased in all patients with a mean of 50 degrees. Pain decreased significantly and none complained of prominence of the proximal ulnar stump in normal pronated position and during a tight grip. The wrist radiographs of both coronal and sagittal planes in normal and stress fisting views were used to evaluate the postoperative static and physiologic loaded stability of the proximal ulnar stump. It had shown this procedure provided good static proximal ulnar stump stability in both coronal and sagittal planes. However, in physiologic loaded condition, it was able to provide stability only in the sagittal plane.

  9. Calibration procedures and first dataset of Southern Ocean chlorophyll a profiles collected by elephant seals equipped with a newly developed CTD-fluorescence tags

    Science.gov (United States)

    Guinet, C.; Xing, X.; Walker, E.; Monestiez, P.; Marchand, S.; Picard, B.; Jaud, T.; Authier, M.; Cotté, C.; Dragon, A. C.; Diamond, E.; Antoine, D.; Lovell, P.; Blain, S.; D'Ortenzio, F.; Claustre, H.

    2013-02-01

    In situ observation of the marine environment has traditionally relied on ship-based platforms. The obvious consequence is that physical and biogeochemical properties have been dramatically undersampled, especially in the remote Southern Ocean (SO). The difficulty in obtaining in situ data represents the major limitations to our understanding, and interpretation of the coupling between physical forcing and the biogeochemical response. Southern elephant seals (Mirounga leonina) equipped with a new generation of oceanographic sensors can measure ocean structure in regions and seasons rarely observed with traditional oceanographic platforms. Over the last few years, seals have allowed for a considerable increase in temperature and salinity profiles from the SO, but we were still lacking information on the spatiotemporal variation of phytoplankton concentration. This information is critical to assess how the biological productivity of the SO, with direct consequences on the amount of CO2 "fixed'' by the biological pump, will respond to global warming. In this research programme, we use an innovative sampling fluorescence approach to quantify phytoplankton concentration at sea. For the first time, a low energy consumption fluorometer was added to Argos CTD-SRDL tags, and these novel instruments were deployed on 27 southern elephant seals between 25 December 2007 and the 4 February 2011. As many as 3388 fluorescence profiles associated with temperature and salinity measurements were thereby collected from a vast sector of the Southern Indian Ocean. This paper addresses the calibration issue of the fluorometer before being deployed on elephant seals and presents the first results obtained for the Indian sector of the Southern Ocean. This in situ system is implemented in synergy with satellite ocean colour radiometry. Satellite-derived data is limited to the surface layer and is restricted over the SO by extensive cloud cover. However, with the addition of these new tags

  10. Radiometric Cross-Calibration of GF-4 in Multispectral Bands

    Directory of Open Access Journals (Sweden)

    Aixia Yang

    2017-03-01

    Full Text Available The GaoFen-4 (GF-4, launched at the end of December 2015, is China’s first high-resolution geostationary optical satellite. A panchromatic and multispectral sensor (PMS is onboard the GF-4 satellite. Unfortunately, the GF-4 has no onboard calibration assembly, so on-orbit radiometric calibration is required. Like the charge-coupled device (CCD onboard HuanJing-1 (HJ or the wide field of view sensor (WFV onboard GaoFen-1 (GF-1, GF-4 also has a wide field of view, which provides challenges for cross-calibration with narrow field of view sensors, like the Landsat series. A new technique has been developed and used to calibrate HJ-1/CCD and GF-1/WFV, which is verified viable. The technique has three key steps: (1 calculate the surface using the bi-directional reflectance distribution function (BRDF characterization of a site, taking advantage of its uniform surface material and natural topographic variation using Landsat Enhanced Thematic Mapper Plus (ETM+/Operational Land Imager (OLI imagery and digital elevation model (DEM products; (2 calculate the radiance at the top-of-the atmosphere (TOA with the simulated surface reflectance using the atmosphere radiant transfer model; and (3 fit the calibration coefficients with the TOA radiance and corresponding Digital Number (DN values of the image. This study attempts to demonstrate the technique is also feasible to calibrate GF-4 multispectral bands. After fitting the calibration coefficients using the technique, extensive validation is conducted by cross-validation using the image pairs of GF-4/PMS and Landsat-8/OLI with similar transit times and close view zenith. The validation result indicates a higher accuracy and frequency than that given by the China Centre for Resources Satellite Data and Application (CRESDA using vicarious calibration. The study shows that the new technique is also quite feasible for GF-4 multispectral bands as a routine long-term procedure.

  11. Site characterization for calibration of radiometric sensors using vicarious method

    Science.gov (United States)

    Parihar, Shailesh; Rathore, L. S.; Mohapatra, M.; Sharma, A. K.; Mitra, A. K.; Bhatla, R.; Singh, R. S.; Desai, Yogdeep; Srivastava, Shailendra S.

    2016-05-01

    Radiometric performances of earth observation satellite/sensors vary from ground pre-launch calibration campaign to post launch period extended to lifetime of the satellite due to launching vibrations. Therefore calibration is carried out worldwide through various methods throughout satellite lifetime. In India Indian Space Research Organization (ISRO) calibrates the sensor of Resourcesat-2 satellite by vicarious method. One of these vicarious calibration methods is the reflectance-based approach that is applied in this study for radiometric calibration of sensors on-board Resouresat-2 satellite. The results of ground-based measurement of atmospheric conditions and surface reflectance are made at Bap, Rajasthan Calibration/Validation (Cal/Val) site. Cal/Val observations at site were carried out with hyper-spectral Spectroradiometer covering spectral range of 350nm- 2500nm for radiometric characterization of the site. The Sunphotometer/Ozonometer for measuring the atmospheric parameters has also been used. The calibrated radiance is converted to absolute at-sensor spectral reflectance and Top-Of-Atmosphere (TOA) radiance. TOA radiance was computed using radiative transfer model `Second simulation of the satellite signal in the solar spectrum' (6S), which can accurately simulate the problems introduced by the presence of the atmosphere along the path from Sun to target (surface) to Sensor. The methodology for band averaged reflectance retrieval and spectral reflectance fitting process are described. Then the spectral reflectance and atmospheric parameters are put into 6S code to predict TOA radiance which compare with Resourcesat-2 radiance. Spectral signature and its reflectance ratio indicate the uniformity of the site. Thus the study proves that the selected site is suitable for vicarious calibration of sensor of Resourcesat-2. Further the study demonstrates the procedure for similar exercise for site selection for Cal/Val analysis of other satellite over India

  12. Congenital pseudoarthrosis of the tibia – results of treatment by free fibular transfer and associated procedures – preliminary study.

    Science.gov (United States)

    Iamaguchi, Raquel B; Fucs, Patricia M M B; Carlos da Costa, Antonio; Chakkour, Ivan; Gomes, Mogar D

    2011-09-01

    We evaluated 16 children with congenital pseudoarthrosis of the tibia treated with contralateral fibular graft, with the aim to report the difficulties and clinical results in the affected limb after consolidation. Sixty-three percent of the children had characteristics of neurofibromatosis. Consolidation was achieved after the main surgery in 37%of patients, and the remainder, after multiple procedures. Consolidation time was longer for male patients. Refracture was observed in six patients and recurrence of the anterior bowing in six; four of these patients were submitted to correction. Four patients presented femur overgrowth. The average shortening of the affected leg was 3.6 cm. The proposed procedure leads to a long treatment course with many reoperations for correction of possible complications.

  13. Normalization and calibration of geostationary satellite radiances for the International Satellite Cloud Climatology Project

    Science.gov (United States)

    Desormeaux, Yves; Rossow, William B.; Brest, Christopher L.; Campbell, G. G.

    1993-01-01

    Procedures are described for normalizing the radiometric calibration of image radiances obtained from geostationary weather satellites that contributed data to the International Satellite Cloud Climatology Project. The key step is comparison of coincident and collocated measurements made by each satellite and the concurrent AVHRR on the 'afternoon' NOAA polar-orbiting weather satellite at the same viewing geometry. The results of this comparison allow transfer of the AVHRR absolute calibration, which has been established over the whole series, to the radiometers on the geostationary satellites. Results are given for Meteosat-2, 3, and 4, for GOES-5, 6, and 7, for GMS-2, 3, and 4 and for Insat-1B. The relative stability of the calibrations of these radiance data is estimated to be within +/- 3 percent; the uncertainty of the absolute calibrations is estimated to be less than 10 percent. The remaining uncertainties are at least two times smaller than for the original radiance data.

  14. Evaluation of the induced electric field and compliance procedure for a wireless power transfer system in an electrical vehicle.

    Science.gov (United States)

    Laakso, Ilkka; Hirata, Akimasa

    2013-11-07

    In this study, an induced electric field in a human body is evaluated for the magnetic field leaked from a wireless power transfer system for charging an electrical vehicle. The magnetic field from the wireless power transfer system is modelled computationally, and its effectiveness is confirmed by comparison with the field measured in a previous study. The induced electric field in a human standing around the vehicle is smaller than the allowable limit prescribed in international guidelines, although the magnetic field strength in the human body is locally higher than the allowable external field strength. Correlation between the external magnetic field and the induced electric field is confirmed to be reasonable at least in the standing posture, which is the case discussed in the international standard. Based on this finding, we discussed and confirmed the applicability of a three-point magnetic field measurement at heights of 0.5, 1.0, and 1.5 m for safety compliance.

  15. Analysis on the revision of the United States authorizing procedure for the transfer of unclassified nuclear technology

    Energy Technology Data Exchange (ETDEWEB)

    Yoon, Sung-ho; Seo, Hana; Lee, Chansuh; Kim, Jong-sook [Korea Institute of Nuclear Nonproliferation and Control, Daejeon (Korea, Republic of)

    2015-10-15

    The DOE (Department Of Energy) has not comprehensively update 10CFR810 since 1986. Since then, the global civil nuclear market has expanded, particularly in China, the Middle East, and Eastern Europe, with vendors from France, Japan, the Republic of Korea, Russia, and Canada. In result, DOE issued revised 810 in respond to comments received from the public and commercial nuclear market changes. This regulation revision improves the efficiency of authorization process to promote national nuclear industry while maintaining nonproliferation control. Even though ROK has initiated a legal basis for Intangible technology transfer (ITT) for nuclear export control, working implementation system is not set up. This research proposes recommendable ITT implementation of the ROK according to the analysis result of the US regulation. In this revision, of 124 countries had been classified as general authorization under 10CFR810, 80 countries reclassified into the specific authorization. By remaining 'fast track' for specific authorization, in particular, time frames for internal DOE and interagency reviews are reduced. This means the US government actively copes with commercial nuclear market expands to promote their industry. Meanwhile, by remaining some of nuclear-weapon states (China, Russia, India) as specific authorization maintaining that the determinations are consistent with current US national security, diplomatic, and trade policy. By benchmarking the US regulation, Korea can improve the efficiency of the technology transfer authorization process easing the regulatory burden by reducing uncertainty and timelines while maintaining the highest level of nonproliferation control.

  16. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  17. Calibration report for ZephIR Dual Mode lidar (unit 351)

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael

    . Their estimation is the ultimate goal of a calibration: a relation is established between reference measurements from calibrated instruments and corresponding LiDAR indications. Traceability in the calibration is obtained by transferring measurement uncertainties from the reference instrument through...

  18. Ørsted Pre-Flight Magnetometer Calibration Mission

    DEFF Research Database (Denmark)

    Risbo, T.; Brauer, Peter; Merayo, José M.G.

    2003-01-01

    The compact spherical coil (CSC) vector-feedback magnetometer on the Danish circle dividersted geomagnetic mapping satellite underwent extensive calibrations and verifications prior to integration and launch. The theory of the 'thin shell' calibration procedure is introduced. Spherical harmonic m...

  19. Multiple View Reconstruction of Calibrated Images using Singular Value Decomposition

    CERN Document Server

    Chaudhury, Ayan; Manna, Sumita; Mukherjee, Subhadeep; Chakrabarti, Amlan

    2010-01-01

    Calibration in a multi camera network has widely been studied for over several years starting from the earlier days of photogrammetry. Many authors have presented several calibration algorithms with their relative advantages and disadvantages. In a stereovision system, multiple view reconstruction is a challenging task. However, the total computational procedure in detail has not been presented before. Here in this work, we are dealing with the problem that, when a world coordinate point is fixed in space, image coordinates of that 3D point vary for different camera positions and orientations. In computer vision aspect, this situation is undesirable. That is, the system has to be designed in such a way that image coordinate of the world coordinate point will be fixed irrespective of the position & orientation of the cameras. We have done it in an elegant fashion. Firstly, camera parameters are calculated in its local coordinate system. Then, we use global coordinate data to transfer all local coordinate d...

  20. Mercury Calibration System

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Joseph Rovani; Mark Sanderson; Ryan Boysen; William Schuster

    2009-03-11

    U.S. Environmental Protection Agency (EPA) Performance Specification 12 in the Clean Air Mercury Rule (CAMR) states that a mercury CEM must be calibrated with National Institute for Standards and Technology (NIST)-traceable standards. In early 2009, a NIST traceable standard for elemental mercury CEM calibration still does not exist. Despite the vacature of CAMR by a Federal appeals court in early 2008, a NIST traceable standard is still needed for whatever regulation is implemented in the future. Thermo Fisher is a major vendor providing complete integrated mercury continuous emissions monitoring (CEM) systems to the industry. WRI is participating with EPA, EPRI, NIST, and Thermo Fisher towards the development of the criteria that will be used in the traceability protocols to be issued by EPA. An initial draft of an elemental mercury calibration traceability protocol was distributed for comment to the participating research groups and vendors on a limited basis in early May 2007. In August 2007, EPA issued an interim traceability protocol for elemental mercury calibrators. Various working drafts of the new interim traceability protocols were distributed in late 2008 and early 2009 to participants in the Mercury Standards Working Committee project. The protocols include sections on qualification and certification. The qualification section describes in general terms tests that must be conducted by the calibrator vendors to demonstrate that their calibration equipment meets the minimum requirements to be established by EPA for use in CAMR monitoring. Variables to be examined include linearity, ambient temperature, back pressure, ambient pressure, line voltage, and effects of shipping. None of the procedures were described in detail in the draft interim documents; however they describe what EPA would like to eventually develop. WRI is providing the data and results to EPA for use in developing revised experimental procedures and realistic acceptance criteria based on

  1. Calibrating cosmological radiative transfer simulations with Lyman alpha forest data: Evidence for large spatial UV background fluctuations at z ~ 5.6 - 5.8 due to rare bright sources

    CERN Document Server

    Chardin, Jonathan; Aubert, Dominique; Puchwein, Ewald

    2015-01-01

    We calibrate here cosmological radiative transfer simulation with ATON/RAMSES with a range of measurements of the Lyman alpha opacity from QSO absorption spectra. We find the Lyman alpha opacity to be very sensitive to the exact timing of hydrogen reionisation. Models reproducing the measured evolution of the mean photoionisation rate and average mean free path reach overlap at z ~ 7 and predict an accelerated evolution of the Lyman alpha opacity at z > 6 consistent with the rapidly evolving luminosity function of Lyman alpha emitters in this redshift range. Similar to "optically thin" simulations our full radiative transfer simulations fail, however, to reproduce the high-opacity tail of the Lyman alpha opacity PDF at z > 5. We argue that this is due to spatial UV fluctuations in the post-overlap phase of reionisation on substantially larger scales than predicted by our source model, where the ionising emissivity is dominated by large numbers of sub-L* galaxies. We further argue that this suggests a signific...

  2. A computational procedure for the investigation of whipping effect on ITER High Energy Piping and its application to the ITER divertor primary heat transfer system

    Energy Technology Data Exchange (ETDEWEB)

    Spagnuolo, G.A., E-mail: Alessandro.Spagnuolo@kraftanlagen.com [Kraftanlagen Heidelberg Gmbh, Im Breitspiel 7, D-69126 Heidelberg (Germany); Dell’Orco, G. [ITER Organization, Route de Vinon-sur-Verdon, CS 90 046, 13067 St. Paul Lez Durance Cedex (France); Di Maio, P.A. [Dipartimento di Energia, Ingegneria dell’Informazione e Modelli Matematici, Università di Palermo Viale delle Scienze, 90128 Palermo (Italy); Mazzei, M. [Kraftanlagen Heidelberg Gmbh, Im Breitspiel 7, D-69126 Heidelberg (Germany)

    2015-10-15

    Highlights: • High Energy Piping (HEP) are components containing water or steam with P ≥ 2.0 MPa and/or T ≥ 100 °C. • The whipping effect in HEP may cause dangerous domino effect with relative rupture propagation. • The rapture is envisaged or postulated according to the stress state of piping. • A FEM analysis is performed in order to study the dynamic of whipping effect. • Study of special support to avoid and/or mitigate the whipping effect. - Abstract: The Tokamak Cooling Water System of nuclear facility has the function to remove heat from plasma facing components maintaining coolant temperatures, pressures and flow rates as required and, depending on thermal-hydraulic requirements, its systems are defined as High Energy Piping (HEP) because they contain fluids, such as water or steam, at a pressure greater than or equal to 2.0 MPa and/or at a temperature greater than or equal to 100 °C, or even gas at pressure above the atmospheric one. The French standards contemplate the need to consider the whipping effect on HEP design. This effect happens when, after a double ended guillotine break, the reaction force could create a displacement of the piping which might affect adjacent components. A research campaign has been performed, in cooperation by ITER Organization and University of Palermo, to outline the procedure to check whether whipping effect might occur and assess its potential damage effects so to allow their mitigation. This procedure is based on the guidelines issued by U.S. Nuclear Regulatory Commission. The proposed procedure has been applied to the analysis of the whipping effect of divertor primary heat transfer system HEP, using a theoretical–computational approach based on the finite element method.

  3. Mexican national pyronometer network calibration

    Science.gov (United States)

    VAldes, M.; Villarreal, L.; Estevez, H.; Riveros, D.

    2013-12-01

    In order to take advantage of the solar radiation as an alternate energy source it is necessary to evaluate the spatial and temporal availability. The Mexican National Meterological Service (SMN) has a network with 136 meteorological stations, each coupled with a pyronometer for measuring the global solar radiation. Some of these stations had not been calibrated in several years. The Mexican Department of Energy (SENER) in order to count on a reliable evaluation of the solar resource funded this project to calibrate the SMN pyrometer network and validate the data. The calibration of the 136 pyronometers by the intercomparison method recommended by the World Meterological Organization (WMO) requires lengthy observations and specific environmental conditions such as clear skies and a stable atmosphere, circumstances that determine the site and season of the calibration. The Solar Radiation Section of the Instituto de Geofísica of the Universidad Nacional Autónoma de México is a Regional Center of the WMO and is certified to carry out the calibration procedures and emit certificates. We are responsible for the recalibration of the pyronometer network of the SMN. A continuous emission solar simulator with exposed areas with 30cm diameters was acquired to reduce the calibration time and not depend on atmospheric conditions. We present the results of the calibration of 10 thermopile pyronometers and one photovoltaic cell by the intercomparison method with more than 10000 observations each and those obtained with the solar simulator.

  4. Geometric calibration of high-resolution remote sensing sensors

    Institute of Scientific and Technical Information of China (English)

    LIANG Hong-you; GU Xing-fa; TAO Yu; QIAO Chao-fei

    2007-01-01

    This paper introduces the applications of high-resolution remote sensing imagery and the necessity of geometric calibration for remote sensing sensors considering assurance of the geometric accuracy of remote sensing imagery. Then the paper analyzes the general methodology of geometric calibration. Taking the DMC sensor geometric calibration as an example, the paper discusses the whole calibration procedure. Finally, it gave some concluding remarks on geometric calibration of high-resolution remote sensing sensors.

  5. TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. Part 2. Dependence on the characteristics of collimator, optical sourse-distance indicator, treatment field, lasers and treatment couch

    Directory of Open Access Journals (Sweden)

    Y. V. Tsitovich

    2016-01-01

    Full Text Available To ensure the safety of radiation oncology patients needed to provide consistent functional characteristics of the medical linear accelerators, which affect the accuracy of dose delivery. To this end, their quality control procedures, which include the calibration of radiation output of the linac, the error in determining the dose reference value during which must not exceed 2 %, is provided. The aim is to develop a methodology for determining the error (difference between a measured value of quantity and its true value in determining this value, depending on the characteristics of the collimator, the source to surface distance pointer, lasers, radiation field and treatment table. To achieve the objectives have been carried out dosimetric measurements of Trilogy S/N 3567 linac dose distributions, on the basis of which dose errors depending on the accuracy setting the zero position of the collimator, the deviation of the collimator rotation isocenter, the sourcesurface distance pointer accuracy, field size accuracy, the accuracy of lasers and treatment table positioning were obtained. It was found that the greatest impact on the value of the error has the error in the optical SSD indication and the error in the lasers position in the plane perpendicular to the plane of incidence of the radiation beam (up to 3.64 % for the energy of 6 MV. Dose errors caused by error in the field size were different for two photon energies, and reached 2.54 % for 6 MeV and 1.33% for 18 MeV. Errors caused by the rest of the characteristic do not exceed 1 %. Thus, it is possible to express the results of periodic quality control of these devices integrated in linac in terms of dose and use them to conduct a comprehensive assessment of the possibility of clinical use of a linear accelerator for oncology patients irradiation on the basis of the calibration of radiation output in case of development of techniques that allow to analyze the influence dosimetric

  6. Does preoperative abduction value affect functional outcome of combined muscle transfer and release procedures in obstetrical palsy patients with shoulder involvement?

    Directory of Open Access Journals (Sweden)

    Onel Defne

    2004-08-01

    Full Text Available Abstract Background Obstetric palsy is the injury of the brachial plexus during delivery. Although many infants with plexopathy recover with minor or no residual functional deficits, some children don't regain sufficient limb function because of functional limitations, bony deformities and joint contractures. Shoulder is the most frequently affected joint with internal rotation contracture causing limitation of abduction, external rotation. The treatment comprises muscle release procedures such as posterior subscapularis sliding or anterior subscapularis tendon lengtening and muscle transfers to restore the missing external rotation and abduction function. Methods We evaluated whether the preoperative abduction degree affects functional outcome. Between 1998 and 2002, 46 children were operated on to restore shoulder abduction and external rotation. The average age at surgery was 7.6 years and average follow up was 40.8 months. We compared the postoperative results of the patients who had preoperative abduction less than 90° (Group I: n = 37 with the patients who had preoperative abduction greater than 90° (Group II: n = 9, in terms of abduction and external rotation function with angle measurements and Mallet classification. We inquired whether patients in Group I needed another muscle transfer along with latissimus dorsi and teres major transfers. Results In Group I the average abduction improved from 62.5° to 131.4° (a 68.9° ± 22.9°gain and the average external rotation improved from 21.4° to 82.6° (a 61.1° ± 23°gain. In Group II the average abduction improved from 99.4°to 140°(a40.5° ± 16°gain and the average external rotation improved from 33.2°to 82.7° (a 49.5° ± 23.9° gain. Although there was a significant difference between Group I and II for preoperative abduction (p = 0.000 and abduction gain in degrees (p = 0.001, the difference between postoperative values of both groups was not significant (p = 0.268. There was

  7. Design and Calibration of a Cryogenic Blackbody Calibrator at Centimeter Wavelengths

    CERN Document Server

    Kogut, A J; Fixsen, D J; Limon, M; Mirel, P G A; Levin, S; Seiffert, M; Lubin, P M

    2004-01-01

    We describe the design and calibration of an external cryogenic blackbody calibrator used for the first two flights of the Absolute Radiometer for Cosmology, Astrophysics, and Diffuse Emission (ARCADE) instrument. The calibrator consists of a microwave absorber weakly coupled to a superfluid liquid helium bath. Half-wave corrugations viewed 30 deg off axis reduce the return loss below -35 dB. Ruthenium oxide resistive thermometers embedded within the absorber monitor the temperature across the face of the calibrator. The thermal calibration transfers the calibration of a reference thermometer to the flight thermometers using the flight thermometer readout system. Data taken near the superfluid transition in 8 independent calibrations 4 years apart agree within 0.3 mK, providing an independent verification of the thermometer calibration at temperatures near that of the cosmic microwave background.

  8. Calibration of ultrasonic sensors of a mobile robot

    Directory of Open Access Journals (Sweden)

    Paunović Ivan

    2009-01-01

    Full Text Available The paper discusses a mobile robot localization. Due to cost and simplicity of signal processing, the ultrasonic sensors are very suitable for this application. However, their nonlinear characteristics requires thorough calibrating procedure in order to achieve reliable readings from the obstacles around the robot. Here we describe SMR400 ultrasonic sensor and its calibration procedure. The suggested calibration procedure was tested through a number of experiments, and the results are presented in this paper. .

  9. Calibration of a single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method

    Energy Technology Data Exchange (ETDEWEB)

    Abbas, Mahmoud I., E-mail: mabbas@physicist.net [Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria (Egypt); Badawi, M.S. [Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria (Egypt); Ruskov, I.N. [Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Institute for Nuclear Research and Nuclear Energy, Bulgarian Academy of Sciences, 1784 Sofia (Bulgaria); El-Khatib, A.M. [Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria (Egypt); Grozdanov, D.N. [Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Institute for Nuclear Research and Nuclear Energy, Bulgarian Academy of Sciences, 1784 Sofia (Bulgaria); Thabet, A.A. [Department of Medical Equipment Technology, Faculty of Allied Medical Sciences, Pharos University in Alexandria (Egypt); Kopatch, Yu.N. [Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Gouda, M.M. [Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria (Egypt); Skoy, V.R. [Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation)

    2015-01-21

    Gamma-ray detector systems are important instruments in a broad range of science and new setup are continually developing. The most recent step in the evolution of detectors for nuclear spectroscopy is the construction of large arrays of detectors of different forms (for example, conical, pentagonal, hexagonal, etc.) and sizes, where the performance and the efficiency can be increased. In this work, a new direct numerical method (NAM), in an integral form and based on the efficiency transfer (ET) method, is used to calculate the full-energy peak efficiency of a single hexagonal NaI(Tl) detector. The algorithms and the calculations of the effective solid angle ratios for a point (isotropic irradiating) gamma-source situated coaxially at different distances from the detector front-end surface, taking into account the attenuation of the gamma-rays in the detector's material, end-cap and the other materials in-between the gamma-source and the detector, are considered as the core of this (ET) method. The calculated full-energy peak efficiency values by the (NAM) are found to be in a good agreement with the measured experimental data.

  10. Sensitivity of Forward Radiative Transfer Model on Spectroscopic Assumptions and Input Geophysical Parameters at 23.8 GHz and 183 GHz Channels and its Impact on Inter-calibration of Microwave Radiometers

    Science.gov (United States)

    Datta, S.; Jones, W. L.; Ebrahimi, H.; Chen, R.; Payne, V.; Kroodsma, R.

    2014-12-01

    The first step in radiometric inter-calibration is to ascertain the self-consistency and reasonableness of the observed brightness temperature (Tb) for each individual sensor involved. One of the widely used approaches is to compare the observed Tb with a simulated Tb using a forward radiative transfer model (RTM) and input geophysical parameters at the geographic location and time of the observation. In this study we intend to test the sensitivity of the RTM to uncertainties in the input geophysical parameters as well as to the underlying physical assumptions of gaseous absorption and surface emission in the RTM. SAPHIR, a cross track scanner onboard Indo-French Megha-Tropique Satellite, gives us a unique opportunity of studying 6 dual band 183 GHz channels at an inclined orbit over the Tropics for the first time. We will also perform the same sensitivity analysis using the Advance Technology Microwave Sounder (ATMS) 23 GHz and five 183 GHz channels. Preliminary analysis comparing GDAS and an independent retrieved profile show some sensitivity of the RTM to the input data. An extended analysis of this work using different input geophysical parameters will be presented. Two different absorption models, the Rosenkranz and the MonoRTM will be tested to analyze the sensitivity of the RTM to spectroscopic assumptions in each model. Also for the 23.8 GHz channel, the sensitivity of the RTM to the surface emissivity model will be checked. Finally the impact of these sensitivities on radiometric inter-calibration of radiometers at sounding frequencies will be assessed.

  11. TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY

    Directory of Open Access Journals (Sweden)

    Y. V. Tsitovich

    2015-01-01

    Full Text Available To ensure the safety of radiation oncology patients needed to provide a consistent functional characteristics of the medical linear electron accelerators, which affect the accuracy of dose delivery. To this end, their quality control procedures, which include the calibration of radiation output of the linear accelerator, the error in determining the dose reference value during which must not exceed 2 %, is provided. The aim is to develop a methodology for determining the error in determining this value, depending on the mechanical charachteristics of the linac’s gantry. To achieve the objectives have been carried out dosimetric measurements of Trilogy S/N 3567 linac dose distributions, on the basis of which dose errors depending on the accuracy setting the zero position of the gantry and the deviation of the gantry rotation isocenter were obtained. It was found that the greatest impact on the value of the error has gantry rotation isocenter deviation in a plane perpendicular to the plane of incidence of the radiation beam (up to 3,64% for the energy of 6 MeV. Dose errors caused by tilting the gantry and its isocenter deviation in the plane of incidence of the beam were highest for 18 MeV energy and reached –0,7 % and –0,9 % respectively. Thus, it is possible to express the results of periodic quality control of the linear accelerator ganty in terms of dose and use them to conduct a comprehensive assessment of the possibility of clinical use of a linear accelerator for oncology patients irradiation in case of development of techniques that allow to analyze the influence of the rest of its technical and dosimetric parameters for error in dose.

  12. Lidar to lidar calibration phase 1

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents a feasibility study of a lidar to lidar (L2L) calibration procedure. Phase one of the project was conducted at Høvsøre, Denmark. Two windcubes were placed next to the 116m met mast and different methods were applied to obtain the sensing height error of the lidars. The purpose...... is to find the most consistent method and use it in a potential lidar to lidar calibration procedure....

  13. Hierarchy of individual calibration levels for heart rate and accelerometry to measure physical activity

    DEFF Research Database (Denmark)

    Brage, Søren; Ekelund, Ulf; Brage, Niels

    2007-01-01

    calibration limits feasibility of these techniques in population studies, and less burdensome, yet valid, methods of calibration are required. We aimed to evaluate the precision of different individual calibration procedures against a reference calibration procedure: a ramped treadmill walking-running test...

  14. Calibration of high resolution remote sensing instruments in the visible and near infrared

    Science.gov (United States)

    Schüller, L.; Fischer, J.; Armbruster, W.; Bartsch, B.

    1997-05-01

    Measurements of the reflected solar radiation with high spectral resolution airborne instruments are usually used to develop new remote sensing techniques. The observed spectral features in the signals provide the possibility to define useful band settings for future satellite instruments. A precise wavelength and radiometric calibration is a prerequisite for such tasks. In this paper, a calibration procedure for the airborne spectrometer OVID is presented. The Optical Visible and near Infrared Detector consists of two similar detector systems, (600 - 1100 nm = VIS and 900 - 1700 nm = NIR). The spectral resolution is ~1.7 nm for the VIS-system and ~6 nm for the IR-system. This instrument is applied for the retrieval of water vapour content, aerosol and cloud properties. Besides the spectral and intensity calibration, also corrections for the dark current signals and for defective pixels have been performed. An indirect verification of the calibration procedure by the comparison of OVID measurements in cloudy and cloud free atmospheres with radiative transfer simulations is discussed in this paper. The used radiation transfer model MOMO is based on the matrix operator method.

  15. On the absolute calibration of SO2 cameras

    Directory of Open Access Journals (Sweden)

    P. Lübcke

    2013-03-01

    Full Text Available Sulphur dioxide emission rate measurements are an important tool for volcanic monitoring and eruption risk assessment. The SO2 camera technique remotely measures volcanic emissions by analysing the ultraviolet absorption of SO2 in a narrow spectral window between 300 and 320 nm using solar radiation scattered in the atmosphere. The SO2 absorption is selectively detected by mounting band-pass interference filters in front of a two-dimensional, UV-sensitive CCD detector. One important step for correct SO2 emission rate measurements that can be compared with other measurement techniques is a correct calibration. This requires conversion from the measured optical density to the desired SO2 column density (CD. The conversion factor is most commonly determined by inserting quartz cells (cuvettes with known amounts of SO2 into the light path. Another calibration method uses an additional narrow field-of-view Differential Optical Absorption Spectroscopy system (NFOV-DOAS, which measures the column density simultaneously in a small area of the camera's field-of-view. This procedure combines the very good spatial and temporal resolution of the SO2 camera technique with the more accurate column densities obtainable from DOAS measurements. This work investigates the uncertainty of results gained through the two commonly used, but quite different, calibration methods (DOAS and calibration cells. Measurements with three different instruments, an SO2 camera, a NFOV-DOAS system and an Imaging DOAS (I-DOAS, are presented. We compare the calibration-cell approach with the calibration from the NFOV-DOAS system. The respective results are compared with measurements from an I-DOAS to verify the calibration curve over the spatial extent of the image. The results show that calibration cells, while working fine in some cases, can lead to an overestimation of the SO2 CD by up to 60% compared with CDs from the DOAS measurements. Besides these errors of calibration

  16. Test surfaces useful for calibration of surface profilometers

    Science.gov (United States)

    Yashchuk, Valeriy V; McKinney, Wayne R; Takacs, Peter Z

    2013-12-31

    The present invention provides for test surfaces and methods for calibration of surface profilometers, including interferometric and atomic force microscopes. Calibration is performed using a specially designed test surface, or the Binary Pseudo-random (BPR) grating (array). Utilizing the BPR grating (array) to measure the power spectral density (PSD) spectrum, the profilometer is calibrated by determining the instrumental modulation transfer.

  17. The VTTVIS line imaging spectrometer - principles, error sources, and calibration

    DEFF Research Database (Denmark)

    Jørgensen, R.N.

    2002-01-01

    work describing the basic principles, potential error sources, and/or adjustment and calibration procedures. This report fulfils the need for such documentationwith special focus on the system at KVL. The PGP based system has several severe error sources, which should be removed prior any analysis....... Most of the random noise sources can be minimised by carefully selecting high-grade components especially withconcern to the camera. Systematic error sources like CCD fixed pattern noise (FPN), CCD photoresponse nonuniformity (PRNU), CCD charge transfer efficiency (CTE), slit width variations, changes...... in off-axis transmission efficiencies, diffractionefficiencies, and image distortion have a significant impact on the instrument performance. Procedures removing or minimising these systematic error sources are developed and described for the system build at KVL but can be generalised to other PGP...

  18. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina

    2016-05-02

    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  19. Calibration of sound calibrators: an overview

    Science.gov (United States)

    Milhomem, T. A. B.; Soares, Z. M. D.

    2016-07-01

    This paper presents an overview of calibration of sound calibrators. Initially, traditional calibration methods are presented. Following, the international standard IEC 60942 is discussed emphasizing parameters, target measurement uncertainty and criteria for conformance to the requirements of the standard. Last, Regional Metrology Organizations comparisons are summarized.

  20. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  1. The Preflight Calibration of the Thermal Infrared Sensor (TIRS) on the Landsat Data Continuity Mission

    Science.gov (United States)

    Smith, Ramsey; Reuter, Dennis; Irons, James; Lunsford, Allen; Montanero, Matthew; Tesfaye, Zelalem; Wenny, Brian; Thome, Kurtis

    2011-01-01

    The preflight calibration testing of TIRS evaluates the performance of the instrument at the component, subsystem and system level, The overall objective is to provide an instrument that is well calibrated and well characterized with specification compliant data that will ensure the data continuity of Landsat from the previous missions to the LDCM, The TIRS flight build unit and the flight instrument were assessed through a series of calibration tests at NASA Goddard Space Flight Center. Instrument-level requirements played a strong role in defining the test equipment and procedures used for the calibration in the thermal/vacuum chamber. The calibration ground support equipment (CGSE), manufactured by MEI and ATK Corporation, was used to measure the optical, radiometric and geometric characteristics of TIRS, The CGSE operates in three test configurations: GeoRad (geometric, radiometric and spatial), flood source and spectral, TIRS was evaluated though the following tests: bright target recovery, radiometry, spectral response, spatial shape, scatter, stray light, focus, and uniformity, Data were obtained for the instrument and various subsystems under conditions simulating those on orbit In the spectral configuration, a monochromator system with a blackbody source is used for in-band and out-of-band relative spectral response characterization, In the flood source configuration the entire focal plane array is illuminated simultaneously to investigate pixel-to-pixel uniformity and dead or inoperable pixels, The remaining tests were executed in the GeoRad configuration and use a NIST calibrated cavity blackbody source, The NIST calibration is transferred to the TIRS sensor and to the blackbody source on-board TIRS, The onboard calibrator will be the primary calibration source for the TIRS sensor on orbit.

  2. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Joseph Rovani; Mark Sanderson

    2008-02-29

    Mercury continuous emissions monitoring systems (CEMS) are being implemented in over 800 coal-fired power plant stacks. The power industry desires to conduct at least a full year of monitoring before the formal monitoring and reporting requirement begins on January 1, 2009. It is important for the industry to have available reliable, turnkey equipment from CEM vendors. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The generators are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 requires that calibration be performed with NIST-traceable standards (Federal Register 2007). Traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued an interim traceability protocol for elemental mercury generators (EPA 2007). The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The document is divided into two separate sections. The first deals with the qualification of generators by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the generator models that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma/mass spectrometry performed by NIST in Gaithersburg, MD. The

  3. On the absolute calibration of SO2 cameras

    Directory of Open Access Journals (Sweden)

    J. Zielcke

    2012-09-01

    Full Text Available Sulphur dioxide emission flux measurements are an important tool for volcanic monitoring and eruption risk assessment. The SO2 camera technique remotely measures volcanic emissions by analysing the ultraviolet absorption of SO2 in a narrow spectral window between 305 nm and 320 nm using solar radiation scattered in the atmosphere. The SO2 absorption is selectively detected by mounting band-pass interference filters in front of a two-dimensional, UV-sensitive CCD detector. While this approach is simple and delivers valuable insights into the two-dimensional SO2 distribution, absolute calibration has proven to be difficult. An accurate calibration of the SO2 camera (i.e., conversion from optical density to SO2 column density, CD is crucial to obtain correct SO2 CDs and flux measurements that are comparable to other measurement techniques and can be used for volcanological applications. The most common approach for calibrating SO2 camera measurements is based on inserting quartz cells (cuvettes containing known amounts of SO2 into the light path. It has been found, however, that reflections from the windows of the calibration cell can considerably affect the signal measured by the camera. Another possibility for calibration relies on performing simultaneous measurements in a small area of the camera's field-of-view (FOV by a narrow-field-of-view Differential Optical Absorption Spectroscopy (NFOV-DOAS system. This procedure combines the very good spatial and temporal resolution of the SO2 camera technique with the more accurate column densities obtainable from DOAS measurements. This work investigates the uncertainty of results gained through the two commonly used, but quite different calibration methods (DOAS and calibration cells. Measurements with three different instruments, an SO2 camera, a NFOV-DOAS system and an Imaging DOAS (IDOAS, are presented. We compare the calibration-cell approach with the calibration from the NFOV-DOAS system. The

  4. GIFTS SM EDU Radiometric and Spectral Calibrations

    Science.gov (United States)

    Tian, J.; Reisse, R. a.; Johnson, D. G.; Gazarik, J. J.

    2007-01-01

    The Geosynchronous Imaging Fourier Transform Spectrometer (GIFTS) Sensor Module (SM) Engineering Demonstration Unit (EDU) is a high resolution spectral imager designed to measure infrared (IR) radiance using a Fourier transform spectrometer (FTS). The GIFTS instrument gathers measurements across the long-wave IR (LWIR), short/mid-wave IR (SMWIR), and visible spectral bands. The raw interferogram measurements are radiometrically and spectrally calibrated to produce radiance spectra, which are further processed to obtain atmospheric profiles via retrieval algorithms. This paper describes the processing algorithms involved in the calibration. The calibration procedures can be subdivided into three categories: the pre-calibration stage, the calibration stage, and finally, the post-calibration stage. Detailed derivations for each stage are presented in this paper.

  5. 40 CFR 86.1524 - Carbon dioxide analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 19 2010-07-01 2010-07-01 false Carbon dioxide analyzer calibration. 86.1524 Section 86.1524 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... Test Procedures § 86.1524 Carbon dioxide analyzer calibration. (a) The calibration requirements for...

  6. 40 CFR 90.425 - CVS calibration frequency.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 20 2010-07-01 2010-07-01 false CVS calibration frequency. 90.425 Section 90.425 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS... Test Procedures § 90.425 CVS calibration frequency. Calibrate the CVS positive displacement pump...

  7. 40 CFR 86.521-90 - Hydrocarbon analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Hydrocarbon analyzer calibration. 86... Regulations for 1978 and Later New Motorcycles; Test Procedures § 86.521-90 Hydrocarbon analyzer calibration. (a) The FID hydrocarbon analyzer shall receive the following initial and periodic calibration....

  8. 40 CFR 1065.325 - Intake-flow calibration.

    Science.gov (United States)

    2010-07-01

    ... CONTROLS ENGINE-TESTING PROCEDURES Calibrations and Verifications Flow-Related Measurements § 1065.325... recommend using a calibration subsonic venturi, ultrasonic flow meter or laminar flow element. We recommend... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Intake-flow calibration....

  9. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values....... Thereafter the code calibration problem is presented in its principal decision theoretical form and it is discussed how acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for ultimate...... and serviceability limit states. Finally the paper describes the Joint Committee on Structural Safety (JCSS) recommended procedure - CodeCal - for the practical implementation of reliability based code calibration of LRFD based design codes....

  10. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  11. Energy calibration issues in nuclear resonant vibrational spectroscopy: observing small spectral shifts and making fast calibrations.

    Science.gov (United States)

    Wang, Hongxin; Yoda, Yoshitaka; Dong, Weibing; Huang, Songping D

    2013-09-01

    The conventional energy calibration for nuclear resonant vibrational spectroscopy (NRVS) is usually long. Meanwhile, taking NRVS samples out of the cryostat increases the chance of sample damage, which makes it impossible to carry out an energy calibration during one NRVS measurement. In this study, by manipulating the 14.4 keV beam through the main measurement chamber without moving out the NRVS sample, two alternative calibration procedures have been proposed and established: (i) an in situ calibration procedure, which measures the main NRVS sample at stage A and the calibration sample at stage B simultaneously, and calibrates the energies for observing extremely small spectral shifts; for example, the 0.3 meV energy shift between the 100%-(57)Fe-enriched [Fe4S4Cl4](=) and 10%-(57)Fe and 90%-(54)Fe labeled [Fe4S4Cl4](=) has been well resolved; (ii) a quick-switching energy calibration procedure, which reduces each calibration time from 3-4 h to about 30 min. Although the quick-switching calibration is not in situ, it is suitable for normal NRVS measurements.

  12. Practical intraoperative stereo camera calibration.

    Science.gov (United States)

    Pratt, Philip; Bergeles, Christos; Darzi, Ara; Yang, Guang-Zhong

    2014-01-01

    Many of the currently available stereo endoscopes employed during minimally invasive surgical procedures have shallow depths of field. Consequently, focus settings are adjusted from time to time in order to achieve the best view of the operative workspace. Invalidating any prior calibration procedure, this presents a significant problem for image guidance applications as they typically rely on the calibrated camera parameters for a variety of geometric tasks, including triangulation, registration and scene reconstruction. While recalibration can be performed intraoperatively, this invariably results in a major disruption to workflow, and can be seen to represent a genuine barrier to the widespread adoption of image guidance technologies. The novel solution described herein constructs a model of the stereo endoscope across the continuum of focus settings, thereby reducing the number of degrees of freedom to one, such that a single view of reference geometry will determine the calibration uniquely. No special hardware or access to proprietary interfaces is required, and the method is ready for evaluation during human cases. A thorough quantitative analysis indicates that the resulting intrinsic and extrinsic parameters lead to calibrations as accurate as those derived from multiple pattern views.

  13. Rapid calibration for line structured light vision sensors

    Institute of Scientific and Technical Information of China (English)

    YANG Pei; XU Bin-shi; WU Lin

    2006-01-01

    The mathematic model of the line structured-light 3-D vision sensor was established.A separating-parameter method was proposed to calibrate the structure parameter,and a calibration target was designed.In the calibrating process,complicated measurement is avoided and the calibration is fast and easy to carry out,which simplifies the calibrating procedure.The experimental results show that the space measurement accuracy is better than 0.15 mm.This method is high efficient and practical in vision sensors calibration.

  14. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  15. The KLOE Online Calibration System

    Institute of Scientific and Technical Information of China (English)

    E.Pasqualucci

    2001-01-01

    Based on all the features of the KLOE online software,the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed.Acalibration manager process controls the system,implementing the interface to the online system,receiving information from the run control and translating its state transitions to a separate state machine.It acts as a " calibration run controller"and performs failure recovery when requested by a set of process checkers.The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms.A client library and C,fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an bool-like interface.Several calibration processes running in parallel in a destributed,multiplatform environment can fill the same histograms,allowing fast external information check.A monitor thread allow remote browsing for visual inspection,Pre-filtered data are read in nonprivileged spy mode from the data acquisition system via the Kloe Integrated Dataflow,privileged spy mode from the data acquisiton system via the Kole Integrated Dataflow.The main characteristics of the system are presented.

  16. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    Science.gov (United States)

    Cornic, Philippe; Illoul, Cédric; Cheminet, Adam; Le Besnerais, Guy; Champagnat, Frédéric; Le Sant, Yves; Leclaire, Benjamin

    2016-09-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data.

  17. Cryogenic Pressure Calibration Facility Using a Cold Force Reference

    CERN Document Server

    Bager, T; Métral, L

    1999-01-01

    Presently various commercial cryogenic pressure sensors are being investigated for installation in the LHC collider, they will eventually be used to assess that the magnets are fully immersed in liquid and to monitor fast pressure transients. In the framework of this selection procedure a cryogenic pressue calibration facility has been designed and built; it is based on a cryogenic primary pressure reference made of a bellows that converts the pressure into a force measurement. For that a shaft transfers this force to a precision force transducer at room temperature. Knowing the liquid bath pessure and the surface area of the bellows the pressure applied to the transducers under calibration is calculated; corrections due to thermal contraction are introduced. To avoid loss of force in the bellows wall its length is maintained constant; a cold capacitive displacement sensor measures this. The calibration temperature covers 1.5 K to 4.2 K and the pressure 0 to 20 bar. In contrast with more classical techniques ...

  18. Development and validation of a fast and simple multi-analyte procedure for quantification of 40 drugs relevant to emergency toxicology using GC-MS and one-point calibration.

    Science.gov (United States)

    Meyer, Golo M J; Weber, Armin A; Maurer, Hans H

    2014-05-01

    Diagnosis and prognosis of poisonings should be confirmed by comprehensive screening and reliable quantification of xenobiotics, for example by gas chromatography-mass spectrometry (GC-MS) or liquid chromatography-mass spectrometry (LC-MS). The turnaround time should be short enough to have an impact on clinical decisions. In emergency toxicology, quantification using full-scan acquisition is preferable because this allows screening and quantification of expected and unexpected drugs in one run. Therefore, a multi-analyte full-scan GC-MS approach was developed and validated with liquid-liquid extraction and one-point calibration for quantification of 40 drugs relevant to emergency toxicology. Validation showed that 36 drugs could be determined quickly, accurately, and reliably in the range of upper therapeutic to toxic concentrations. Daily one-point calibration with calibrators stored for up to four weeks reduced workload and turn-around time to less than 1 h. In summary, the multi-analyte approach with simple liquid-liquid extraction, GC-MS identification, and quantification over fast one-point calibration could successfully be applied to proficiency tests and real case samples.

  19. Thermal conductivity calibration for hot wire based dc scanning thermal microscopy

    Science.gov (United States)

    Lefèvre, Stéphane; Volz, Sebastian; Saulnier, Jean-Bernard; Fuentes, Catherine; Trannoy, Nathalie

    2003-04-01

    Thermal conductivity characterization with nanoscale spatial resolution can be performed by contact probe techniques only. The technique based on a hot anemometer wire probe mounted in an atomic force microscope is now a standard setup. However, no rigorous calibration procedure is provided so far in basic dc mode. While in contact with the sample surface, the electrical current I injected into the probe is controlled so that electrical resistance or the wire temperature is maintained by the Joule effect. The variation in current is assumed to be linearly related to the heat flux lost towards the sample and traditional calibration is carried out by relating the thermal conductivity of a set of samples to the measured current I. We provide analytical and numerical thermal modeling of the tip and sample to estimate the key heat transfer in a conductivity calibration procedure. A simple calibration expression is established that provides thermal conductivity as a function of the probe current or voltage measured. Finally, experimental data allow us to determine the unknown quantities of the parametric form obtained, i.e., the mean tip-sample contact radius and conductance.

  20. The ATLAS Electromagnetic Calorimeter Calibration Workshop

    CERN Multimedia

    Hong Ma; Isabelle Wingerter

    The ATLAS Electromagnetic Calorimeter Calibration Workshop took place at LAPP-Annecy from the 1st to the 3rd of October; 45 people attended the workshop. A detailed program was setup before the workshop. The agenda was organised around very focused presentations where questions were raised to allow arguments to be exchanged and answers to be proposed. The main topics were: Electronics calibration Handling of problematic channels Cluster level corrections for electrons and photons Absolute energy scale Streams for calibration samples Calibration constants processing Learning from commissioning Forty-five people attended the workshop. The workshop was on the whole lively and fruitful. Based on years of experience with test beam analysis and Monte Carlo simulation, and the recent operation of the detector in the commissioning, the methods to calibrate the electromagnetic calorimeter are well known. Some of the procedures are being exercised in the commisssioning, which have demonstrated the c...

  1. Trinocular Calibration Method Based on Binocular Calibration

    Directory of Open Access Journals (Sweden)

    CAO Dan-Dan

    2012-10-01

    Full Text Available In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error of the global calibration of the two camera pairs in the experiments can be as low as 0.277% and 0.328% respectively. Experiment results show that this method is feasible, simple and effective, and has high precision.

  2. Calibration of hydrological model with programme PEST

    Science.gov (United States)

    Brilly, Mitja; Vidmar, Andrej; Kryžanowski, Andrej; Bezak, Nejc; Šraj, Mojca

    2016-04-01

    PEST is tool based on minimization of an objective function related to the root mean square error between the model output and the measurement. We use "singular value decomposition", section of the PEST control file, and Tikhonov regularization method for successfully estimation of model parameters. The PEST sometimes failed if inverse problems were ill-posed, but (SVD) ensures that PEST maintains numerical stability. The choice of the initial guess for the initial parameter values is an important issue in the PEST and need expert knowledge. The flexible nature of the PEST software and its ability to be applied to whole catchments at once give results of calibration performed extremely well across high number of sub catchments. Use of parallel computing version of PEST called BeoPEST was successfully useful to speed up calibration process. BeoPEST employs smart slaves and point-to-point communications to transfer data between the master and slaves computers. The HBV-light model is a simple multi-tank-type model for simulating precipitation-runoff. It is conceptual balance model of catchment hydrology which simulates discharge using rainfall, temperature and estimates of potential evaporation. Version of HBV-light-CLI allows the user to run HBV-light from the command line. Input and results files are in XML form. This allows to easily connecting it with other applications such as pre and post-processing utilities and PEST itself. The procedure was applied on hydrological model of Savinja catchment (1852 km2) and consists of twenty one sub-catchments. Data are temporary processed on hourly basis.

  3. Calibration for 3D Structured Light Measurement

    Institute of Scientific and Technical Information of China (English)

    2007-01-01

    A calibration procedure was developed for three-dimensional(3D) binocular structured light measurement systems. In virtue of a specially designed pattern, matching points in stereo images are extracted. And then sufficient 3D space points are obtained through pairs of images with the intrinsic and extrinsic parameters of each camera estimated prior and consequently some lights are calibrated by means of multi point fitting. Finally, a mathematical model is applied to interpolate and approximate all dynamic scanning lights based on geometry. The process of calibration method is successfully used in the binocular 3D measurement system based on structured lights and the 3D reconstruction results are satisfying.

  4. Increased Automation in Stereo Camera Calibration Techniques

    Directory of Open Access Journals (Sweden)

    Brandi House

    2006-08-01

    Full Text Available Robotic vision has become a very popular field in recent years due to the numerous promising applications it may enhance. However, errors within the cameras and in their perception of their environment can cause applications in robotics to fail. To help correct these internal and external imperfections, stereo camera calibrations are performed. There are currently many accurate methods of camera calibration available; however, most or all of them are time consuming and labor intensive. This research seeks to automate the most labor intensive aspects of a popular calibration technique developed by Jean-Yves Bouguet. His process requires manual selection of the extreme corners of a checkerboard pattern. The modified process uses embedded LEDs in the checkerboard pattern to act as active fiducials. Images are captured of the checkerboard with the LEDs on and off in rapid succession. The difference of the two images automatically highlights the location of the four extreme corners, and these corner locations take the place of the manual selections. With this modification to the calibration routine, upwards of eighty mouse clicks are eliminated per stereo calibration. Preliminary test results indicate that accuracy is not substantially affected by the modified procedure. Improved automation to camera calibration procedures may finally penetrate the barriers to the use of calibration in practice.

  5. Generic methodology for calibrating profiling nacelle lidars

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael; Wagner, Rozenn

    Improving power performance assessment by measuring at different heights has been demonstrated using ground-based profiling LIDARs. More recently, nacelle-mounted lidars studies have shown promising capabilities to assess power performance. Using nacelle lidars avoids the erection of expensive...... meteorology masts, especially offshore. A new generation of commercially developed profiling nacelle lidars has sophisticated measurement capabilities. As for any other measuring system, lidars measurements have uncertainties. Their estimation is the ultimate goal of a calibration. Field calibration...... procedures have been developed for non-profiling nacelle lidars. However, their specificity to one type of lidar or another highlights the need for developing generic calibration procedures. Such procedures should be applicable to any type of existing and upcoming lidar technology. Profiling nacelle lidars...

  6. An actor-oriented transfer perspective on high school students’ development of the use of procedures to solve problems on “rate of change”.

    NARCIS (Netherlands)

    Roorda, Gerrit; Goedhart, Martin; Vos, Pauline

    2015-01-01

    This article reports on a longitudinal observation study about students’ development in their use of procedures to calculate instantaneous rate of change. Different procedures for solving tasks on rate of change are taught in mathematics and physics classes, and together they form a repertoire. Our

  7. An Actor-Oriented Transfer Perspective on High School Students' Development of the Use of Procedures to Solve Problems on Rate of Change

    Science.gov (United States)

    Roorda, Gerrit; Vos, Pauline; Goedhart, Martin J.

    2015-01-01

    This article reports on a longitudinal observation study about students' development in their use of procedures to calculate instantaneous rate of change. Different procedures for solving tasks on rate of change are taught in mathematics and physics classes, and together they form a repertoire. Our study took an actor-oriented perspective, which…

  8. Calibration of the Hydrological Simulation Program Fortran (HSPF) model using automatic calibration and geographical information systems

    Science.gov (United States)

    Al-Abed, N. A.; Whiteley, H. R.

    2002-11-01

    Calibrating a comprehensive, multi-parameter conceptual hydrological model, such as the Hydrological Simulation Program Fortran model, is a major challenge. This paper describes calibration procedures for water-quantity parameters of the HSPF version 10·11 using the automatic-calibration parameter estimator model coupled with a geographical information system (GIS) approach for spatially averaged properties. The study area was the Grand River watershed, located in southern Ontario, Canada, between 79° 30 and 80° 57W longitude and 42° 51 and 44° 31N latitude. The drainage area is 6965 km2. Calibration efforts were directed to those model parameters that produced large changes in model response during sensitivity tests run prior to undertaking calibration. A GIS was used extensively in this study. It was first used in the watershed segmentation process. During calibration, the GIS data were used to establish realistic starting values for the surface and subsurface zone parameters LZSN, UZSN, COVER, and INFILT and physically reasonable ratios of these parameters among watersheds were preserved during calibration with the ratios based on the known properties of the subwatersheds determined using GIS. This calibration procedure produced very satisfactory results; the percentage difference between the simulated and the measured yearly discharge ranged between 4 to 16%, which is classified as good to very good calibration. The average simulated daily discharge for the watershed outlet at Brantford for the years 1981-85 was 67 m3 s-1 and the average measured discharge at Brantford was 70 m3 s-1. The coupling of a GIS with automatice calibration produced a realistic and accurate calibration for the HSPF model with much less effort and subjectivity than would be required for unassisted calibration.

  9. Evaluation of an ASM1 Model Calibration Precedure on a Municipal-Industrial Wastewater Treatment Plant

    DEFF Research Database (Denmark)

    Petersen, Britta; Gernaey, Krist; Henze, Mogens;

    2002-01-01

    The purpose of the calibrated model determines how to approach a model calibration, e.g. which information is needed and to which level of detail the model should be calibrated. A systematic model calibration procedure was therefore defined and evaluated for a municipal–industrial wastewater trea...

  10. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    Science.gov (United States)

    Jessen, W.; Wilbert, S.; Nouri, B.; Geuder, N.; Fritz, H.

    2015-10-01

    Resource assessment for Concentrated Solar Power (CSP) needs accurate Direct Normal Irradiance (DNI) measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs) with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA) in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May) show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  11. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    Directory of Open Access Journals (Sweden)

    W. Jessen

    2015-10-01

    Full Text Available Resource assessment for Concentrated Solar Power (CSP needs accurate Direct Normal Irradiance (DNI measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  12. Study of Lever-Arm Effect Using Embedded Photogrammetry and On-Board GPS Receiver on Uav for Metrological Mapping Purpose and Proposal of a Free Ground Measurements Calibration Procedure

    Science.gov (United States)

    Daakir, M.; Pierrot-Deseilligny, M.; Bosser, P.; Pichard, F.; Thom, C.; Rabot, Y.

    2016-03-01

    Nowadays, Unmanned Aerial Vehicle (UAV) on-board photogrammetry knows a significant growth due to the democratization of using drones in the civilian sector. Also, due to changes in regulations laws governing the rules of inclusion of a UAV in the airspace which become suitable for the development of professional activities. Fields of application of photogrammetry are diverse, for instance: architecture, geology, archaeology, mapping, industrial metrology, etc. Our research concerns the latter area. Vinci-Construction- Terrassement is a private company specialized in public earthworks that uses UAVs for metrology applications. This article deals with maximum accuracy one can achieve with a coupled camera and GPS receiver system for direct-georeferencing of Digital Surface Models (DSMs) without relying on Ground Control Points (GCPs) measurements. This article focuses specially on the lever-arm calibration part. This proposed calibration method is based on two steps: a first step involves the proper calibration for each sensor, i.e. to determine the position of the optical center of the camera and the GPS antenna phase center in a local coordinate system relative to the sensor. A second step concerns a 3d modeling of the UAV with embedded sensors through a photogrammetric acquisition. Processing this acquisition allows to determine the value of the lever-arm offset without using GCPs.

  13. Innovative methodology for intercomparison of radionuclide calibrators using short half-life in situ prepared radioactive sources

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, P. A. [Centro de Investigação do Instituto Português de Oncologia do Porto Francisco Gentil, EPE, Porto, Portugal and Departamento de Física e Astronomia, Faculdade de Ciências da Universidade do Porto (Portugal); Santos, J. A. M., E-mail: joao.santos@ipoporto.min-saude.pt [Centro de Investigação do Instituto Português de Oncologia do Porto Francisco Gentil, EPE, Porto (Portugal); Serviço de Física Médica do Instituto Português de Oncologia do Porto Francisco Gentil, EPE, Porto (Portugal); Serviço de Medicina Nuclear do Instituto Português de Oncologia do Porto Francisco Gentil, EPE, Porto (Portugal); Instituto de Ciências Biomédicas Abel Salazar, Universidade do Porto, Porto (Portugal)

    2014-07-15

    Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as{sup 99m}Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a careful analysis of the methodology, for the case of{sup 99m}Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator. Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source preparedin situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using {sup 99m}Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results. Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for{sup 99m}Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half-life radionuclides.

  14. Sky camera geometric calibration using solar observations

    Science.gov (United States)

    Urquhart, Bryan; Kurtz, Ben; Kleissl, Jan

    2016-09-01

    A camera model and associated automated calibration procedure for stationary daytime sky imaging cameras is presented. The specific modeling and calibration needs are motivated by remotely deployed cameras used to forecast solar power production where cameras point skyward and use 180° fisheye lenses. Sun position in the sky and on the image plane provides a simple and automated approach to calibration; special equipment or calibration patterns are not required. Sun position in the sky is modeled using a solar position algorithm (requiring latitude, longitude, altitude and time as inputs). Sun position on the image plane is detected using a simple image processing algorithm. The performance evaluation focuses on the calibration of a camera employing a fisheye lens with an equisolid angle projection, but the camera model is general enough to treat most fixed focal length, central, dioptric camera systems with a photo objective lens. Calibration errors scale with the noise level of the sun position measurement in the image plane, but the calibration is robust across a large range of noise in the sun position. Calibration performance on clear days ranged from 0.94 to 1.24 pixels root mean square error.

  15. Local Volatility Calibration Using An Adjoint Proxy

    Directory of Open Access Journals (Sweden)

    Gabriel TURINICI

    2008-11-01

    Full Text Available We document the calibration of the local volatility in a framework similar to Coleman, Li and Verma. The quality of a surface is assessed through a functional to be optimized; the specificity of the approach is to separate the optimization (performed with any suitable optimization algorithm from the computation of the functional where we use an adjoint (as in L. Jiang et. al. to obtain an approximation; moreover our main calibration variable is the implied volatility (the procedure can also accommodate the Greeks. The procedure performs well on benchmarks from the literature and on FOREX data.

  16. The calibration of PIXIE

    Science.gov (United States)

    Fixsen, D. J.; Chuss, D. T.; Kogut, Alan; Mirel, Paul; Wollack, E. J.

    2016-07-01

    The FIRAS instrument demonstrated the use of an external calibrator to compare the sky to an instrumented blackbody. The PIXIE calibrator is improved from -35 dB to -65 dB. Another significant improvement is the ability to insert the calibrator into either input of the FTS. This allows detection and correction of additional errors, reduces the effective calibration noise by a factor of 2, eliminates an entire class of systematics and allows continuous observations. This paper presents the design and use of the PIXIE calibrator.

  17. Multiple height calibration artefact for 3D microscopy

    DEFF Research Database (Denmark)

    De Chiffre, Leonardo; Carli, Lorenzo; Eriksen, Rasmus Solmer

    2011-01-01

    A novel artefact for calibration of the height in 3D microscopy is presented. The artefact comprises three steps having a common vertical axis, which allows z-coordinate calibration at different magnifications without requiring repositioning. The artefact is suitable for transferring traceability...

  18. Regularities of heat transfer in the gas layers of a steam boiler furnace flame. Part II. Gas layer radiation laws and the procedure for calculating heat transfer in furnaces, fire boxes, and combustion chambers developed on the basis of these laws

    Science.gov (United States)

    Makarov, A. N.

    2014-10-01

    The article presents the results stemming from the scientific discovery of laws relating to radiation from the gas layers generated during flame combustion of fuel and when electric arc burns in electric-arc steel-melting furnaces. The procedure for calculating heat transfer in electric-arc and torch furnaces, fire-boxes, and combustion chambers elaborated on the basis of this discovery is described.

  19. Distributed Radio Interferometric Calibration

    CERN Document Server

    Yatawatta, Sarod

    2015-01-01

    Increasing data volumes delivered by a new generation of radio interferometers require computationally efficient and robust calibration algorithms. In this paper, we propose distributed calibration as a way of improving both computational cost as well as robustness in calibration. We exploit the data parallelism across frequency that is inherent in radio astronomical observations that are recorded as multiple channels at different frequencies. Moreover, we also exploit the smoothness of the variation of calibration parameters across frequency. Data parallelism enables us to distribute the computing load across a network of compute agents. Smoothness in frequency enables us reformulate calibration as a consensus optimization problem. With this formulation, we enable flow of information between compute agents calibrating data at different frequencies, without actually passing the data, and thereby improving robustness. We present simulation results to show the feasibility as well as the advantages of distribute...

  20. Calibrators and control samples for bilirubinometers.

    Science.gov (United States)

    Blijenberg, B G; Brügmann, G; Geilenkeuser, W J; Kusyschyn, R; Röhle, G; Schlebusch, H; Schneider, C

    1993-06-01

    The different matrix properties of neonatal serum and commercial control samples can lead to considerable errors in the calibration and control of bilirubinometers. These difficulties can be avoided by calibration with serum from healthy adults which is supplemented with unconjugated bilirubin. But this procedure is impracticable for most routine laboratories. Under certain preconditions, control samples, with bilirubin concentrations determined with correctly calibrated bilirubinometers or spectrophotometers, are also suitable as calibrators. This was established by determination of the bilirubin concentration of 16 different control samples, using both the reference method and correctly calibrated bilirubinometers or spectrophotometers in three or four specialist laboratories. This was also confirmed in several interlaboratory surveys, some involving up to 72 laboratories. The results of these investigations show that a control sample should be used for the calibration of a bilirubinometer only if it meets the following preconditions: 1. There should be no significant difference between the bilirubin values determined with the reference method and with a correctly calibrated spectrophotometer or bilirubinometer. 2. The bilirubin concentration should lie in the range 230-300 mumol/l. The photometric response of bilirubinometers has a limited linear range, so that analytical results greater than 300 mumol/l must be rated as basically unreliable.

  1. Standard guide for establishing calibration for a measurement method used to analyze nuclear fuel cycle materials

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2003-01-01

    1.1 This guide provides the basis for establishing calibration for a measurement method typically used in an analytical chemistry laboratory analyzing nuclear materials. Guidance is included for such activities as preparing a calibration procedure, selecting a calibration standard, controlling calibrated equipment, and documenting calibration. The guide is generic and any required technical information specific for a given method must be obtained from other sources.

  2. Calibration and equivalency analysis of image plate scanners

    Energy Technology Data Exchange (ETDEWEB)

    Williams, G. Jackson, E-mail: williams270@llnl.gov; Maddox, Brian R.; Chen, Hui [Lawrence Livermore National Laboratory, 7000 East Avenue, Livermore, California 94550 (United States); Kojima, Sadaoki [Institute of Laser Engineering, Osaka University, Yamada-oka, 2-6, Suita, Osaka 565-0871 (Japan); Millecchia, Matthew [Laboratory for Laser Energetics, University of Rochester, 250 East River Road, Rochester, New York 14623 (United States)

    2014-11-15

    A universal procedure was developed to calibrate image plate scanners using radioisotope sources. Techniques to calibrate scanners and sources, as well as cross-calibrate scanner models, are described to convert image plate dosage into physical units. This allows for the direct comparison of quantitative data between any facility and scanner. An empirical relation was also derived to establish sensitivity response settings for arbitrary gain settings. In practice, these methods may be extended to any image plate scanning system.

  3. Developmental competence of equine oocytes and embryos obtained by in vitro procedures ranging from in vitro maturation and ICSI to embryo culture, cryopreservation and somatic cell nuclear transfer.

    Science.gov (United States)

    Galli, C; Colleoni, S; Duchi, R; Lagutina, I; Lazzari, G

    2007-03-01

    Development of assisted reproductive technologies in horses has been relatively slow compared to other domestic species, namely ruminants and pigs. The scarce availability of abattoir ovaries and the lack of interest from horse breeders and breed associations have been the main reasons for this delay. Progressively though, the technology of oocyte maturation in vitro has been established followed by the application of ICSI to achieve fertilization in vitro. Embryo culture was initially performed in vivo, in the mare oviduct or in the surrogate sheep oviduct, to achieve the highest embryo development, in the range of 18-36% of the fertilised oocytes. Subsequently, the parallel improvement of in vitro oocyte maturation conditions and embryo culture media has permitted high rates of embryo development from in vitro matured and in vitro cultured ICSI embryos, ranging from 5 to 10% in the early studies to up to 38% in the latest ones. From 2003, with the birth of the first cloned equids, the technology of somatic cell nuclear transfer has also become established due to improvement of the basic steps of embryo production in vitro, including cryopreservation. Pregnancy and foaling rates are still estimated based on a small number of in vitro produced equine embryos transferred to recipients. The largest set of data on non-surgical embryo transfer of in vitro produced embryos, from ICSI of both abattoir and in vitro-matured Ovum Pick Up (OPU) oocytes, and from somatic cell nuclear transfer, has been obtained in our laboratory. The data demonstrate that equine embryos produced by OPU and then cryopreserved can achieve up to 69% pregnancy rate with a foaling rate of 83%. These percentages are reduced to 11 and 23%, respectively, for cloned embryos. In conclusion, extensive evidence exists that in vitro matured equine oocytes can efficiently develop into viable embryos and offspring.

  4. Suomi NPP VIIRS Reflective Solar Bands Operational Calibration Reprocessing

    Directory of Open Access Journals (Sweden)

    Slawomir Blonski

    2015-12-01

    Full Text Available Radiometric calibration coefficients for the VIIRS (Visible Infrared Imaging Radiometer Suite reflective solar bands have been reprocessed from the beginning of the Suomi NPP (National Polar-orbiting Partnership mission until present. An automated calibration procedure, implemented in the NOAA (National Oceanic and Atmospheric Administration JPSS (Joint Polar Satellite System operational data production system, was applied to reprocess onboard solar calibration data and solar diffuser degradation measurements. The latest processing parameters from the operational system were used to include corrected solar vectors, optimized directional dependence of attenuation screens transmittance and solar diffuser reflectance, updated prelaunch calibration coefficients without an offset term, and optimized Robust Holt-Winters filter parameters. The parameters were consistently used to generate a complete set of the radiometric calibration coefficients for the entire duration of the Suomi NPP mission. The reprocessing has demonstrated that the automated calibration procedure can be successfully applied to all solar measurements acquired from the beginning of the mission until the full deployment of the automated procedure in the operational processing system. The reprocessed calibration coefficients can be further used to reprocess VIIRS SDR (Sensor Data Record and other data products. The reprocessing has also demonstrated how the automated calibration procedure can be used during activation of the VIIRS instruments on the future JPSS satellites.

  5. Flux calibration of the Herschel(star)-SPIRE photometer

    NARCIS (Netherlands)

    Bendo, G. J.; Griffin, M. J.; Bock, J. J.; Conversi, L.; Dowell, C. D.; Lim, T.; Lu, N.; North, C. E.; Papageorgiou, A.; Pearson, C. P.; Pohlen, M.; Polehampton, E. T.; Schulz, B.; Shupe, D. L.; Sibthorpe, B.; Spencer, L. D.; Swinyard, B. M.; Valtchanov, I.; Xu, C. K.

    2013-01-01

    We describe the procedure used to flux calibrate the three-band submillimetre photometer in the Spectral and Photometric Imaging Receiver instrument on the Herschel Space Observatory. This includes the equations describing the calibration scheme, a justification for using Neptune as the primary cali

  6. 40 CFR 86.1321-94 - Hydrocarbon analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 19 2010-07-01 2010-07-01 false Hydrocarbon analyzer calibration. 86... Procedures § 86.1321-94 Hydrocarbon analyzer calibration. The FID hydrocarbon analyzer shall receive the... into service and at least annually thereafter, the FID hydrocarbon analyzer shall be adjusted...

  7. 40 CFR 86.121-90 - Hydrocarbon analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Hydrocarbon analyzer calibration. 86... Complete Heavy-Duty Vehicles; Test Procedures § 86.121-90 Hydrocarbon analyzer calibration. The hydrocarbon... FID and HFID hydrocarbon analyzers shall be adjusted for optimum hydrocarbon response....

  8. 40 CFR 1065.315 - Pressure, temperature, and dewpoint calibration.

    Science.gov (United States)

    2010-07-01

    ...) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Calibrations and Verifications Measurement of Engine...) Pressure. We recommend temperature-compensated, digital-pneumatic, or deadweight pressure calibrators, with... quantities that are NIST-traceable within 0.5% uncertainty. (2) Temperature. We recommend digital...

  9. 40 CFR 86.516-90 - Calibrations, frequency and overview.

    Science.gov (United States)

    2010-07-01

    ... PROGRAMS (CONTINUED) CONTROL OF EMISSIONS FROM NEW AND IN-USE HIGHWAY VEHICLES AND ENGINES Emission Regulations for 1978 and Later New Motorcycles; Test Procedures § 86.516-90 Calibrations, frequency and... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Calibrations, frequency and...

  10. 40 CFR 86.226-94 - Calibration of other equipment.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Calibration of other equipment. 86.226-94 Section 86.226-94 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... New Medium-Duty Passenger Vehicles; Cold Temperature Test Procedures § 86.226-94 Calibration of...

  11. Calibration and Temperature Profile of a Tungsten Filament Lamp

    Science.gov (United States)

    de Izarra, Charles; Gitton, Jean-Michel

    2010-01-01

    The goal of this work proposed for undergraduate students and teachers is the calibration of a tungsten filament lamp from electric measurements that are both simple and precise, allowing to determine the temperature of tungsten filament as a function of the current intensity. This calibration procedure was first applied to a conventional filament…

  12. A calibration protocol of a one-dimensional moving bed bioreactor (MBBR) dynamic model for nitrogen removal.

    Science.gov (United States)

    Barry, U; Choubert, J-M; Canler, J-P; Héduit, A; Robin, L; Lessard, P

    2012-01-01

    This work suggests a procedure to correctly calibrate the parameters of a one-dimensional MBBR dynamic model in nitrification treatment. The study deals with the MBBR configuration with two reactors in series, one for carbon treatment and the other for nitrogen treatment. Because of the influence of the first reactor on the second one, the approach needs a specific calibration strategy. Firstly, a comparison between measured values and simulated ones obtained with default parameters has been carried out. Simulated values of filtered COD, NH(4)-N and dissolved oxygen are underestimated and nitrates are overestimated compared with observed data. Thus, nitrifying rate and oxygen transfer into the biofilm are overvalued. Secondly, a sensitivity analysis was carried out for parameters and for COD fractionation. It revealed three classes of sensitive parameters: physical, diffusional and kinetic. Then a calibration protocol of the MBBR dynamic model was proposed. It was successfully tested on data recorded at a pilot-scale plant and a calibrated set of values was obtained for four parameters: the maximum biofilm thickness, the detachment rate, the maximum autotrophic growth rate and the oxygen transfer rate.

  13. Assessment of target volume doses in radiotherapy based on the standard and measured calibration curves

    Directory of Open Access Journals (Sweden)

    Gholamreza Fallah Mohammadi

    2015-01-01

    Full Text Available Context: In radiation treatments, estimation of the dose distribution in the target volume is one of the main components of the treatment planning procedure. To estimate the dose distribution, the information of electron densities is necessary. The standard curves determined by computed tomography (CT scanner that may be different from that of other oncology centers. In this study, the changes of dose calculation due to the different calibration curves (HU-ρel were investigated. Materials and Methods: Dose values were calculated based on the standard calibration curve that was predefined for the treatment planning system (TPS. The calibration curve was also extracted from the CT images of the phantom, and dose values were calculated based on this curve. The percentage errors of the calculated values were determined. Statistical Analysis Used: The statistical analyses of the mean differences were performed using the Wilcoxon rank-sum test for both of the calibration curves. Results and Discussion: The results show no significant difference for both of the measured and standard calibration curves (HU-ρel in 6, 15, and 18 MeV energies. In Wilcoxon ranked sum nonparametric test for independent samples with P < 0.05, the equality of monitor units for both of the curves to transfer 200 cGy doses to reference points was resulted. The percentage errors of the calculated values were lower than 2% and 1.5% in 6 and 15 MeV, respectively. Conclusion: From the results, it could be concluded that the standard calibration curve could be used in TPS dose calculation accurately.

  14. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Fernandez Garcia, Sergio; Villanueva, Héctor

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and correspondi...

  15. Exploration of new multivariate spectral calibration algorithms.

    Energy Technology Data Exchange (ETDEWEB)

    Van Benthem, Mark Hilary; Haaland, David Michael; Melgaard, David Kennett; Martin, Laura Elizabeth; Wehlburg, Christine Marie; Pell, Randy J. (The Dow Chemical Company, Midland, MI); Guenard, Robert D. (Merck & Co. Inc., West Point, PA)

    2004-03-01

    A variety of multivariate calibration algorithms for quantitative spectral analyses were investigated and compared, and new algorithms were developed in the course of this Laboratory Directed Research and Development project. We were able to demonstrate the ability of the hybrid classical least squares/partial least squares (CLSIPLS) calibration algorithms to maintain calibrations in the presence of spectrometer drift and to transfer calibrations between spectrometers from the same or different manufacturers. These methods were found to be as good or better in prediction ability as the commonly used partial least squares (PLS) method. We also present the theory for an entirely new class of algorithms labeled augmented classical least squares (ACLS) methods. New factor selection methods are developed and described for the ACLS algorithms. These factor selection methods are demonstrated using near-infrared spectra collected from a system of dilute aqueous solutions. The ACLS algorithm is also shown to provide improved ease of use and better prediction ability than PLS when transferring calibrations between near-infrared calibrations from the same manufacturer. Finally, simulations incorporating either ideal or realistic errors in the spectra were used to compare the prediction abilities of the new ACLS algorithm with that of PLS. We found that in the presence of realistic errors with non-uniform spectral error variance across spectral channels or with spectral errors correlated between frequency channels, ACLS methods generally out-performed the more commonly used PLS method. These results demonstrate the need for realistic error structure in simulations when the prediction abilities of various algorithms are compared. The combination of equal or superior prediction ability and the ease of use of the ACLS algorithms make the new ACLS methods the preferred algorithms to use for multivariate spectral calibrations.

  16. Sandia WIPP calibration traceability

    Energy Technology Data Exchange (ETDEWEB)

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  17. WFPC2 Polarization Calibration

    Science.gov (United States)

    Biretta, J.; McMaster, M.

    1997-12-01

    We derive a detailed calibration for WFPC2 polarization data which is accurate to about 1.5%. We begin by computing polarizer flats, and show how they are applied to data. A physical model for the polarization effects of the WFPC2 optics is then created using Mueller matricies. This model includes corrections for the instrumental polarization (diattenuation and phase retardance) of the pick-off mirror, as well as the high cross-polarization transmission of the polarizer filter. We compare this model against the on-orbit observations of polarization calibrators, and show it predicts relative counts in the different polarizer/aperture settings to 1.5% RMS accuracy. We then show how this model can be used to calibrate GO data, and present two WWW tools which allow observers to easily calibrate their data. Detailed examples are given illustrationg the calibration and display of WFPC2 polarization data. In closing we describe future plans and possible improvements.

  18. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities; Metodos de calibracao e de intercomparacao de calibradores de dose utilizados em servicos de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Alessandro Martins da

    1999-07-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  19. Uniform calibration of night vision goggles and test sets

    Science.gov (United States)

    Eppeldauer, George P.

    2007-10-01

    There are orders of magnitude differences between the ~0.1 % (k=2) uncertainty of NIST reference detector calibrations and the uncertainty of night vision (NV) goggle measurements. NIST developed a night vision radiometer calibration facility including NV radiometer transfer standards. The transfer standards, that propagate the radiance responsivity scale to the military primary standards laboratories, are calibrated against a NIST reference radiometer. The reference radiometer has been calibrated on the NIST Spectral Comparator Facility (SCF) for spectral power and irradiance responsivities. Spectral considerations are discussed to lower the uncertainties of the radiance responsivity scale transfer to the test sets and then to the goggles. Since direct determination of the final uncertainties in goggle calibrations and measurements is difficult, models have been made to estimate the most important uncertainty components based on individual spectral measurements of the applied source distributions and radiometer spectral responsivities. It is also shown, that because of source spectral mismatch problems, the goggle measurement uncertainty at applications can be much higher than at calibration. A suggestion is being made to mimic the no-moon (stars only) night sky radiation distribution using several LEDs in the test-sets to decrease the large spectral mismatch errors. A broad-band correction factor has been developed to further decrease calibration uncertainty when the goggles to be used have different spectral responsivities than the standard. Geometrical considerations to optimize the radiance measurement angle and the out-of-target blocking are also discussed to decrease the uncertainty in the radiance responsivity transfer.

  20. Segment Based Camera Calibration

    Institute of Scientific and Technical Information of China (English)

    马颂德; 魏国庆; 等

    1993-01-01

    The basic idea of calibrating a camera system in previous approaches is to determine camera parmeters by using a set of known 3D points as calibration reference.In this paper,we present a method of camera calibration in whih camera parameters are determined by a set of 3D lines.A set of constraints is derived on camea parameters in terms of perspective line mapping.Form these constraints,the same perspective transformation matrix as that for point mapping can be computed linearly.The minimum number of calibration lines is 6.This result generalizes that of Liu,Huang and Faugeras[12] for camera location determination in which at least 8 line correspondences are required for linear computation of camera location.Since line segments in an image can be located easily and more accurately than points,the use of lines as calibration reference tends to ease the computation in inage preprocessing and to improve calibration accuracy.Experimental results on the calibration along with stereo reconstruction are reported.

  1. Monte Carlo based calibration of scintillation detectors for laboratory and in situ gamma ray measurements

    NARCIS (Netherlands)

    van der Graaf, E. R.; Limburg, J.; Koomans, R. L.; Tijs, M.

    2011-01-01

    The calibration of scintillation detectors for gamma radiation in a well characterized setup can be transferred to other geometries using Monte Carlo simulations to account for the differences between the calibration and the other geometry. In this study a calibration facility was used that is const

  2. Time Calibration of the ANTARES Neutrino Telescope

    CERN Document Server

    Aguilar, J A; Albert, A; André, M; Anghinolfi, M; Anton, G; Anvar, S; Ardid, M; Jesus, A C Assis; Astraatmadja, T; Aubert, J J; Auer, R; Baret, B; Basa, S; Bazzotti, M; Bertin, V; Biagi, S; Bigongiari, C; Bou-Cabo, M; Bouwhuis, M C; Brown, A M; Brunner, J; Busto, J; Camarena, F; Capone, A; Carloganu, C; Carminati, G; Carr, J; Cecchini, S; Charvis, Ph; Chiarusi, T; Circella, M; Costantini, H; Cottini, N; Coyle, P; Curtil, C; Decowski, M P; Dekeyser, I; Deschamps, A; Distefano, C; Donzaud, C; Dornic, D; Drouhin, D; Eberl, T; Emanuele, U; Ernenwein, J P; Escoffier, S; Fehr, F; Flaminio, V; Fritsch, U; Fuda, J L; Galata, S; Gay, P; Giacomelli, G; Gómez-González, J P; Graf, K; Guillard, G; Halladjian, G; Hallewell, G; van Haren, H; Heijboer, A J; Hello, Y; Hernández-Rey, J J; Herold, B; Hössl, J; Hsu, C C; de Jong, M; Kadler, M; Kalantar-Nayestanaki, N; Kalekin, O; Kappes, A; Katz, U; Kooijman, P; Kopper, C; Kouchner, A; Kulikovskiy, V; Lahmann, R; Lamare, P; Larosa, G; Lefèvre, D; Lim, G; Presti, D Lo; Loehner, H; Loucatos, S; Lucarelli, F; Mangano, S; Marcelin, M; Margiotta, A; Martinez-Mora, J A; Mazure, A; Montaruli, T; Morganti, M; Moscoso, L; Motz, H; Naumann, C; Neff, M; Palioselitis, D; Pavalas, G E; Payre, P; Petrovic, J; Piattelli, P; Picot-Clemente, N; Picq, C; Popa, V; Pradier, T; Presani, E; Racca, C; Reed, C; Riccobene, G; Richardt, C; Rujoiu, M; Russo, G V; Salesa, F; Sapienza, P; Schöck, F; Schuller, J P; Shanidze, R; Simeone, F; Spies, A; Spurio, M; Steijger, J J M; Stolarczyk, Th; Taiuti, M; Tamburini, C; Tasca, L; Toscano, S; Vallage, B; Van Elewyck, V; Vannoni, G; Vecchi, M; Vernin, P; Wijnker, G; de Wolf, E; Yepes, H; Zaborov, D; Zornoza, J D; Zúñiga, J

    2010-01-01

    The ANTARES deep-sea neutrino telescope comprises a three-dimensional array of photomultipliers to detect the Cherenkov light induced by upgoing relativistic charged particles originating from neutrino interactions in the vicinity of the detector. The large scattering length of light in the deep sea facilitates an angular resolution of a few tenths of a degree for neutrino energies exceeding 10 TeV. In order to achieve this optimal performance, the time calibration procedures should ensure a relative time calibration between the photomultipliers at the level of about 1ns. The methods developed to attain this level of precision are described.

  3. Time calibration of the ANTARES neutrino telescope

    Science.gov (United States)

    ANTARES Collaboration; Aguilar, J. A.; Al Samarai, I.; Albert, A.; André, M.; Anghinolfi, M.; Anton, G.; Anvar, S.; Ardid, M.; Assis Jesus, A. C.; Astraatmadja, T.; Aubert, J. J.; Auer, R.; Baret, B.; Basa, S.; Bazzotti, M.; Bertin, V.; Biagi, S.; Bigongiari, C.; Bou-Cabo, M.; Bouwhuis, M. C.; Brown, A. M.; Brunner, J.; Busto, J.; Camarena, F.; Capone, A.; Cârloganu, C.; Carminati, G.; Carr, J.; Cecchini, S.; Charvis, Ph.; Chiarusi, T.; Circella, M.; Costantini, H.; Cottini, N.; Coyle, P.; Curtil, C.; Decowski, M. P.; Dekeyser, I.; Deschamps, A.; Distefano, C.; Donzaud, C.; Dornic, D.; Drouhin, D.; Eberl, T.; Emanuele, U.; Ernenwein, J. P.; Escoffier, S.; Fehr, F.; Flaminio, V.; Fritsch, U.; Fuda, J. L.; Galata, S.; Gay, P.; Giacomelli, G.; Gómez-González, J. P.; Graf, K.; Guillard, G.; Halladjian, G.; Hallewell, G.; van Haren, H.; Heijboer, A. J.; Hello, Y.; Hernández-Rey, J. J.; Herold, B.; Hößl, J.; Hsu, C. C.; de Jong, M.; Kadler, M.; Kalantar-Nayestanaki, N.; Kalekin, O.; Kappes, A.; Katz, U.; Kooijman, P.; Kopper, C.; Kouchner, A.; Kulikovskiy, V.; Lahmann, R.; Lamare, P.; Larosa, G.; Lefèvre, D.; Lim, G.; Lo Presti, D.; Loehner, H.; Loucatos, S.; Lucarelli, F.; Mangano, S.; Marcelin, M.; Margiotta, A.; Martinez-Mora, J. A.; Mazure, A.; Montaruli, T.; Morganti, M.; Moscoso, L.; Motz, H.; Naumann, C.; Neff, M.; Palioselitis, D.; Păvălaş, G. E.; Payre, P.; Petrovic, J.; Piattelli, P.; Picot-Clemente, N.; Picq, C.; Popa, V.; Pradier, T.; Presani, E.; Racca, C.; Reed, C.; Riccobene, G.; Richardt, C.; Rujoiu, M.; Russo, G. V.; Salesa, F.; Sapienzap, P.; Schöck, F.; Schuller, J. P.; Shanidze, R.; Simeone, F.; Spies, A.; Spurio, M.; Steijger, J. J. M.; Stolarczyk, Th.; Taiuti, M.; Tamburini, C.; Tasca, L.; Toscano, S.; Vallage, B.; van Elewyck, V.; Vannoni, G.; Vecchi, M.; Vernin, P.; Wijnker, G.; de Wolf, E.; Yepes, H.; Zaborov, D.; Zornoza, J. D.; Zúñiga, J.; ANTARES Collaboration

    2011-02-01

    The ANTARES deep-sea neutrino telescope comprises a three-dimensional array of photomultipliers to detect the Cherenkov light induced by upgoing relativistic charged particles originating from neutrino interactions in the vicinity of the detector. The large scattering length of light in the deep sea facilitates an angular resolution of a few tenths of a degree for neutrino energies exceeding 10 TeV. In order to achieve this optimal performance, the time calibration procedures should ensure a relative time calibration between the photomultipliers at the level of ˜1 ns. The methods developed to attain this level of precision are described.

  4. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Georgieva Yankova, Ginka; Courtney, Michael

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and corresponding...... lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from the reference lidar measurements are given for information only....

  5. Site Calibration report

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Vesth, Allan

    This report describes the site calibration carried out at Østerild, during a given period. The site calibration was performed with two Windcube WLS7 (v1) lidars at ten measurements heights. The lidar is not a sensor approved by the current version of the IEC 61400-12-1 [1] and therefore the site...... calibration with lidars does not comply with the standard. However, the measurements are carried out following the guidelines of IEC 61400-12-1 where possible, but with some deviations presented in the following chapters....

  6. Calibration Fixture For Anemometer Probes

    Science.gov (United States)

    Lewis, Charles R.; Nagel, Robert T.

    1993-01-01

    Fixture facilitates calibration of three-dimensional sideflow thermal anemometer probes. With fixture, probe oriented at number of angles throughout its design range. Readings calibrated as function of orientation in airflow. Calibration repeatable and verifiable.

  7. Radioactive standards and calibration methods for contamination monitoring instruments

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida, Makoto [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    1997-06-01

    Contamination monitoring in the facilities for handling unsealed radioactive materials is one of the most important procedures for radiation protection as well as radiation dose monitoring. For implementation of the proper contamination monitoring, radiation measuring instruments should not only be suitable to the purpose of monitoring, but also be well calibrated for the objective qualities of measurement. In the calibration of contamination monitoring instruments, quality reference activities need to be used. They are supplied in different such as extended sources, radioactive solutions or radioactive gases. These reference activities must be traceable to the national standards or equivalent standards. On the other hand, the appropriate calibration methods must be applied for each type of contamination monitoring instruments. In this paper, the concepts of calibration for contamination monitoring instruments, reference sources, determination methods of reference quantities and practical calibration methods of contamination monitoring instruments, including the procedures carried out in Japan Atomic Energy Research Institute and some relevant experimental data. (G.K.)

  8. Skew redundant MEMS IMU calibration using a Kalman filter

    Science.gov (United States)

    Jafari, M.; Sahebjameyan, M.; Moshiri, B.; Najafabadi, T. A.

    2015-10-01

    In this paper, a novel calibration procedure for skew redundant inertial measurement units (SRIMUs) based on micro-electro mechanical systems (MEMS) is proposed. A general model of the SRIMU measurements is derived which contains the effects of bias, scale factor error and misalignments. For more accuracy, the effect of lever arms of the accelerometers to the center of the table are modeled and compensated in the calibration procedure. Two separate Kalman filters (KFs) are proposed to perform the estimation of error parameters for gyroscopes and accelerometers. The predictive error minimization (PEM) stochastic modeling method is used to simultaneously model the effect of bias instability and random walk noise on the calibration Kalman filters to diminish the biased estimations. The proposed procedure is simulated numerically and has expected experimental results. The calibration maneuvers are applied using a two-axis angle turntable in a way that the persistency of excitation (PE) condition for parameter estimation is met. For this purpose, a trapezoidal calibration profile is utilized to excite different deterministic error parameters of the accelerometers and a pulse profile is used for the gyroscopes. Furthermore, to evaluate the performance of the proposed KF calibration method, a conventional least squares (LS) calibration procedure is derived for the SRIMUs and the simulation and experimental results compare the functionality of the two proposed methods with each other.

  9. Air Data Calibration Facility

    Data.gov (United States)

    Federal Laboratory Consortium — This facility is for low altitude subsonic altimeter system calibrations of air vehicles. Mission is a direct support of the AFFTC mission. Postflight data merge is...

  10. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina; Webb, Craig

    2016-05-02

    This presentation provides a high-level overview of the progress on the Broadband Outdoor Radiometer Calibrations for all shortwave and longwave radiometers that are deployed by the Atmospheric Radiation Measurement program.

  11. SRHA calibration curve

    Data.gov (United States)

    U.S. Environmental Protection Agency — an UV calibration curve for SRHA quantitation This dataset is associated with the following publication: Chang, X., and D. Bouchard. Surfactant-Wrapped Multiwalled...

  12. TWSTFT Link Calibration Report

    Science.gov (United States)

    2015-09-01

    box calibrator with unknown but constant total delay during a calibration tour Total Delay: The total electrical delay from the antenna phase center...to the UTCp including all the devices/cables that the satellite and clock signals pass through. It numerically equals the sum of all the sub-delays...PTB. To average out the dimnal effects and measurement noise , 5-7 days of continuous measurements is required. 3 Setups at the Lab(k) The setup

  13. Chang'E Microwave Radiometer Data Calibration with LRO Diviner Data and Machine Learning

    Science.gov (United States)

    Tsang, Ken; Hu, Guo-Ping; Zheng, Yong-Chun; This work is supported by BNU-HKBU United International College Research Grant R201626, Zhuhai Premier Discipline Enhancement Grant code: R1050, and Science and Technology Development Fund in Macao SAR 039/2013/A2

    2016-10-01

    Following usual practice in microwave remote sensing, raw data from multi-channel microwave radiometers (MR) onboard the Chinese Chang'E lunar obiters (CE1 & CE2) were acquired as observed antenna voltages, which were then calibrated and converted to brightness temperatures (TB) by a two-point calibration procedure. While the CE cold calibration antenna is supposed to point to the deep space and taking data for the cold reference point in the two-point calibration scheme, in reality, it picked up undesirable thermal microwave radiation from the lunar surface. Thus the "cold" reference point is not exactly the 2.7K cosmic background assumed and this affects the quality of the calibration.In this work, the small but puzzling differences between the two sets of Level 2C MR data released for CE1 & 2 are attributed to the difference in orbital altitudes between CE1 & 2. This leads to the different degrees of contamination to the cold antenna on CE1 & 2 by thermal radiations from the lunar surface, which showed up as persistent lower night-time TB values in the Level 2C CE2 dataset.We proposed a machine learning approach applied directly to pre-Level 2C data in the voltages to TB convertion process. Since all the antenna voltage data as well as the high temperature referencing point in the calibration procedure are directly measurable, optimized regression algorithms have been employed to determine the effective low temperature referencing points and obtain a single set of statistical consistent TB by combining raw data from CE1 & 2, due to the fact that seasonal variations are less than resolution of the CE MR data from low to medium latitudes.Finally, the Lunar Reconnaissance Orbiter (LRO) Diviner IR data are used as constraints on the boundary condition of the top layer regolith temperature to obtain a consistent sub-surface temperature profile, from which the measured CE MR data can be computed through multi-layer radiation transfer model. This step removes most of

  14. Energy calibration via correlation

    CERN Document Server

    Maier, Daniel

    2015-01-01

    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241 Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be le...

  15. The automatic calibration of Korean VLBI Network data

    CERN Document Server

    Hodgson, Jeffrey A; Zhao, Guang-Yao; Algaba, Juan-Carlos; Yun, Youngjoo; Jung, Taehyun; Byun, Do-Young

    2016-01-01

    The calibration of Very Long Baseline Interferometry (VLBI) data has long been a time consuming process. The Korean VLBI Network (KVN) is a simple array consisting of three identical antennas. Because four frequencies are observed simultaneously, phase solutions can be transferred from lower frequencies to higher frequencies in order to improve phase coherence and hence sensitivity at higher frequencies. Due to the homogeneous nature of the array, the KVN is also well suited for automatic calibration. In this paper we describe the automatic calibration of single-polarisation KVN data using the KVN Pipeline and comparing the results against VLBI data that has been manually reduced. We find that the pipelined data using phase transfer produces better results than a manually reduced dataset not using the phase transfer. Additionally we compared the pipeline results with a manually reduced phase-transferred dataset and found the results to be identical.

  16. The Automatic Calibration of Korean VLBI Network Data

    Science.gov (United States)

    Hodgson, Jeffrey A.; Lee, Sang-Sung; Zhao, Guang-Yao; Algaba, Juan-Carlos; Yun, Youngjoo; Jung, Taehyun; Byun, Do-Young

    2016-08-01

    The calibration of Very Long Baseline Interferometry (VLBI) data has long been a time consuming process. The Korean VLBI Network (KVN) is a simple array consisting of three identical antennas. Because four frequencies are observed simultaneously, phase solutions can be transferred from lower frequencies to higher frequencies in order to improve phase coherence and hence sensitivity at higher frequencies. Due to the homogeneous nature of the array, the KVN is also well suited for automatic calibration. In this paper we describe the automatic calibration of single-polarisation KVN data using the KVN Pipeline and comparing the results against VLBI data that has been manually reduced. We find that the pipelined data using phase transfer produces better results than a manually reduced dataset not using the phase transfer. Additionally we compared the pipeline results with a manually reduced phase-transferred dataset and found the results to be identical.

  17. Simple method for calibrating omnidirectional stereo with multiple cameras

    Science.gov (United States)

    Ha, Jong-Eun; Choi, I.-Sak

    2011-04-01

    Cameras can give useful information for the autonomous navigation of a mobile robot. Typically, one or two cameras are used for this task. Recently, an omnidirectional stereo vision system that can cover the whole surrounding environment of a mobile robot is adopted. They usually adopt a mirror that cannot offer uniform spatial resolution. In this paper, we deal with an omnidirectional stereo system which consists of eight cameras where each two vertical cameras constitute one stereo system. Camera calibration is the first necessary step to obtain 3D information. Calibration using a planar pattern requires many images acquired under different poses so it is a tedious step to calibrate all eight cameras. In this paper, we present a simple calibration procedure using a cubic-type calibration structure that surrounds the omnidirectional stereo system. We can calibrate all the cameras on an omnidirectional stereo system in just one shot.

  18. 40 CFR 92.117 - Gas meter or flow instrumentation calibration, particulate measurement.

    Science.gov (United States)

    2010-07-01

    ... calibration, particulate measurement. 92.117 Section 92.117 Protection of Environment ENVIRONMENTAL PROTECTION... ENGINES Test Procedures § 92.117 Gas meter or flow instrumentation calibration, particulate measurement... orifice, a bellmouth nozzle, or a laminar flow element or an NIST traceable flow calibration device...

  19. Calibrating transport lines using LOCO techniques

    Energy Technology Data Exchange (ETDEWEB)

    Yves Roblin

    2011-09-01

    With the 12GeV upgrade underway at CEBAF, there is a need to re-characterize the beamlines after the modifications made to it to accommodate running at higher energies. We present a linear perturbation approach to calibrating the optics model of transport lines. This method is adapted from the LOCO method in use for storage rings. We consider the effect of quadrupole errors, dipole construction errors as well as beam position monitors and correctors calibrations. The ideal model is expanded to first order in Taylor series of the quadrupole errors. A set of difference orbits obtained by exciting the correctors along the beamline is taken, yielding the measured response matrix. An iterative procedure is invoked and the quadrupole errors as well as beam position monitors and corrector calibration factors are obtained. Here we present details of the method and results of first measurements at CEBAF in early 2011.

  20. Quantum-chemistry based calibration of the alkali metal cation series (Li(+)-Cs(+)) for large-scale polarizable molecular mechanics/dynamics simulations.

    Science.gov (United States)

    Dudev, Todor; Devereux, Mike; Meuwly, Markus; Lim, Carmay; Piquemal, Jean-Philip; Gresh, Nohad

    2015-02-15

    The alkali metal cations in the series Li(+)-Cs(+) act as major partners in a diversity of biological processes and in bioinorganic chemistry. In this article, we present the results of their calibration in the context of the SIBFA polarizable molecular mechanics/dynamics procedure. It relies on quantum-chemistry (QC) energy-decomposition analyses of their monoligated complexes with representative O-, N-, S-, and Se- ligands, performed with the aug-cc-pVTZ(-f) basis set at the Hartree-Fock level. Close agreement with QC is obtained for each individual contribution, even though the calibration involves only a limited set of cation-specific parameters. This agreement is preserved in tests on polyligated complexes with four and six O- ligands, water and formamide, indicating the transferability of the procedure. Preliminary extensions to density functional theory calculations are reported.

  1. Calibration Standards for Surface Topography Measuring Systems down to Nanometric Range

    DEFF Research Database (Denmark)

    Trumpold, H.; De Chiffre, Leonardo; Andreasen, Jan Lasson

    calibrator has been designed, built and tested. Primary calibration standards have been produced by ion-beam and plasma etching (step height standards), by holographic generation of sinusoidal structures with two-beam interference exposure and by ultra-precision diamond cutting. From primary standards...... to be replicated during all stages of the replication processes. Procedures for cleaning glass, PVC, PC, PMM and Ni-surfaces have been developed and tested. Calibration procedures for calibration standards and for calibrating instruments in X-, Y- and Z-direction have been developed and tested. Proposals...

  2. 核电项目建造阶段中的焊接工艺评定转移%Welding procedure qualification transference to different project of nuclear power item during the construction period

    Institute of Scientific and Technical Information of China (English)

    马新朝

    2011-01-01

    论述了核电站在建造阶段焊接工艺评定报告由已完成建造任务的核电项目至正在建设中的核电项目中的焊接工艺评定报告的转移情况,分析了工艺评定转移的可行性,转移后施工单位、监理单位、政府监督单位所应关注的重点,突出了质量保证及质量控制,更关注了政府监督单位(核安全局及地区核与辐射监督站)的监督职能.焊接工艺评定的有效转移,一方面节约了重复评定时的人力和物力投入,降低了建造成本;另一方面有力的推动了工程的顺利建设,促进焊接技术及经验的持续发展,从而提离了管理水平.%The paper shows us the Welding producer report transference to different project of nuclear power item during the construction period,analyses the practicability to the producer report transference. After transference,the paper discusses the focal point to welding in construction enterprises,surveillance management corporation as well as the responsible for supervision departments of the Government,give prominence to the Quality assurance and Quality control. The Nuclear Safety Bureau and the Area Departments Responsible for Supervision on radiation should give more attention to the surveillance on transference.The effective transfer of PQR can reduce expenses ,material and manpower in duplicate welding procedure qualification,reduce construction costs ;the other side,which can strongly promotes the project construction,accelerate the welding technology and experience continue development,and improve the management standards.

  3. Absolute calibration of optical power for PDT: report of AAPM TG140.

    Science.gov (United States)

    Zhu, Timothy C; Bonnerup, Chris; Colussi, Valdir C; Dowell, Marla L; Finlay, Jarod C; Lilge, Lothar; Slowey, Thomas W; Sibata, Claudio

    2013-08-01

    This report is primarily concerned with methods for optical calibration of laser power for continuous wave (CW) light sources, predominantly used in photodynamic therapy (PDT). Light power calibration is very important for PDT, however, no clear standard has been established for the calibration procedure nor the requirements of power meters suitable for optical power calibration. The purposes of the report are to provide guidance for establishing calibration procedures for thermopile type power meters and establish calibration uncertainties for most commercially available detectors and readout assemblies. The authors have also provided a review of the use of various power meters for CW and pulsed optical sources, and provided recommended temporal frequencies for optical power meter calibrations and guidance for routine quality assurance procedure.

  4. Adaptable Multivariate Calibration Models for Spectral Applications

    Energy Technology Data Exchange (ETDEWEB)

    THOMAS,EDWARD V.

    1999-12-20

    Multivariate calibration techniques have been used in a wide variety of spectroscopic situations. In many of these situations spectral variation can be partitioned into meaningful classes. For example, suppose that multiple spectra are obtained from each of a number of different objects wherein the level of the analyte of interest varies within each object over time. In such situations the total spectral variation observed across all measurements has two distinct general sources of variation: intra-object and inter-object. One might want to develop a global multivariate calibration model that predicts the analyte of interest accurately both within and across objects, including new objects not involved in developing the calibration model. However, this goal might be hard to realize if the inter-object spectral variation is complex and difficult to model. If the intra-object spectral variation is consistent across objects, an effective alternative approach might be to develop a generic intra-object model that can be adapted to each object separately. This paper contains recommendations for experimental protocols and data analysis in such situations. The approach is illustrated with an example involving the noninvasive measurement of glucose using near-infrared reflectance spectroscopy. Extensions to calibration maintenance and calibration transfer are discussed.

  5. Landsat-8 Operational Land Imager Radiometric Calibration and Stability

    Directory of Open Access Journals (Sweden)

    Brian Markham

    2014-12-01

    Full Text Available The Landsat-8 Operational Land Imager (OLI was radiometrically calibrated prior to launch in terms of spectral radiance, using an integrating sphere source traceable to National Institute of Standards and Technology (NIST standards of spectral irradiance. It was calibrated on-orbit in terms of reflectance using diffusers characterized prior to launch using NIST traceable standards. The radiance calibration was performed with an uncertainty of ~3%; the reflectance calibration to an uncertainty of ~2%. On-orbit, multiple calibration techniques indicate that the sensor has been stable to better than 0.3% to date, with the exception of the shortest wavelength band, which has degraded about 1.0%. A transfer to orbit experiment conducted using the OLI’s heliostat-illuminated diffuser suggests that some bands increased in sensitivity on transition to orbit by as much as 5%, with an uncertainty of ~2.5%. On-orbit comparisons to other instruments and vicarious calibration techniques show the radiance (without a transfer to orbit adjustment, and reflectance calibrations generally agree with other instruments and ground measurements to within the uncertainties. Calibration coefficients are provided with the data products to convert to either radiance or reflectance units.

  6. Euromet action 428: transfer of ge detectors efficiency calibration from point source geometry to other geometries; Action euromet 428: transfert de l'etalonnage en rendement de detecteurs au germanium pour une source ponctuelle vers d'autres geometries

    Energy Technology Data Exchange (ETDEWEB)

    Lepy, M.Ch

    2000-07-01

    The EUROMET project 428 examines efficiency transfer computation for Ge gamma-ray spectrometers when the efficiency is known for a reference point source geometry in the 60 keV to 2 MeV energy range. For this, different methods are used, such as Monte Carlo simulation or semi-empirical computation. The exercise compares the application of these methods to the same selected experimental cases to determine the usage limitations versus the requested accuracy. For carefully examining these results and trying to derive information for improving the computation codes, this study was limited to a few simple cases, from an experimental efficiency calibration for point source at 10-cm source-to-detector distance. The first part concerns the simplest case of geometry transfer, i.e., using point sources for 3 source-to-detector distances: 2,5 and 20 cm; the second part deals with transfer from point source geometry to cylindrical geometry with three different matrices. The general results show that the deviations between the computed results and the measured efficiencies are for the most part within 10%. The quality of the results is rather inhomogeneous and shows that these codes cannot be used directly for metrological purposes. However, most of them are operational for routine measurements when efficiency uncertainties of 5-10% can be sufficient. (author)

  7. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  8. HAWC Timing Calibration

    CERN Document Server

    Huentemeyer, Petra; Dingus, Brenda

    2009-01-01

    The High-Altitude Water Cherenkov (HAWC) Experiment is a second-generation highsensitivity gamma-ray and cosmic-ray detector that builds on the experience and technology of the Milagro observatory. Like Milagro, HAWC utilizes the water Cherenkov technique to measure extensive air showers. Instead of a pond filled with water (as in Milagro) an array of closely packed water tanks is used. The event direction will be reconstructed using the times when the PMTs in each tank are triggered. Therefore, the timing calibration will be crucial for reaching an angular resolution as low as 0.25 degrees.We propose to use a laser calibration system, patterned after the calibration system in Milagro. Like Milagro, the HAWC optical calibration system will use ~1 ns laser light pulses. Unlike Milagro, the PMTs are optically isolated and require their own optical fiber calibration. For HAWC the laser light pulses will be directed through a series of optical fan-outs and fibers to illuminate the PMTs in approximately one half o...

  9. Microhotplate Temperature Sensor Calibration and BIST.

    Science.gov (United States)

    Afridi, M; Montgomery, C; Cooper-Balis, E; Semancik, S; Kreider, K G; Geist, J

    2011-01-01

    In this paper we describe a novel long-term microhotplate temperature sensor calibration technique suitable for Built-In Self Test (BIST). The microhotplate thermal resistance (thermal efficiency) and the thermal voltage from an integrated platinum-rhodium thermocouple were calibrated against a freshly calibrated four-wire polysilicon microhotplate-heater temperature sensor (heater) that is not stable over long periods of time when exposed to higher temperatures. To stress the microhotplate, its temperature was raised to around 400 °C and held there for days. The heater was then recalibrated as a temperature sensor, and microhotplate temperature measurements were made based on the fresh calibration of the heater, the first calibration of the heater, the microhotplate thermal resistance, and the thermocouple voltage. This procedure was repeated 10 times over a period of 80 days. The results show that the heater calibration drifted substantially during the period of the test while the microhotplate thermal resistance and the thermocouple-voltage remained stable to within about plus or minus 1 °C over the same period. Therefore, the combination of a microhotplate heater-temperature sensor and either the microhotplate thermal resistance or an integrated thin film platinum-rhodium thermocouple can be used to provide a stable, calibrated, microhotplate-temperature sensor, and the combination of the three sensor is suitable for implementing BIST functionality. Alternatively, if a stable microhotplate-heater temperature sensor is available, such as a properly annealed platinum heater-temperature sensor, then the thermal resistance of the microhotplate and the electrical resistance of the platinum heater will be sufficient to implement BIST. It is also shown that aluminum- and polysilicon-based temperature sensors, which are not stable enough for measuring high microhotplate temperatures (>220 °C) without impractically frequent recalibration, can be used to measure the

  10. Calibration of line structured light vision system based on camera's projective center

    Institute of Scientific and Technical Information of China (English)

    ZHU Ji-gui; LI Yan-jun; YE Sheng-hua

    2005-01-01

    Based on the characteristics of line structured light sensor, a speedy method for the calibration was established. With the coplanar reference target, the spacial pose between camera and optical plane can be calibrated by using of the camera's projective center and the light's information in the camera's image surface. Without striction to the movement of the coplanar reference target and assistant adjustment equipment, this calibration method can be implemented. This method has been used and decreased the cost of calibration equipment, simplified the calibration procedure, improved calibration efficiency. Using experiment, the sensor can attain relative accuracy about 0.5%, which indicates the rationality and effectivity of this method.

  11. Spectra Transfer Between a Fourier Transform Near-Infrared Laboratory and a Miniaturized Handheld Near-Infrared Spectrometer.

    Science.gov (United States)

    Hoffmann, Uwe; Pfeifer, Frank; Hsuing, Chang; Siesler, Heinz W

    2016-05-01

    The aim of this contribution is to demonstrate the transfer of spectra that have been measured on two different laboratory Fourier transform near-infrared (FT-NIR) spectrometers to the format of a handheld instrument by measuring only a few samples with both spectrometer types. Thus, despite the extreme differences in spectral range and resolution, spectral data sets that have been collected and quantitative as well as qualitative calibrations that have been developed thereof, respectively, over a long period on a laboratory instrument can be conveniently transferred to the handheld system. Thus, the necessity to prepare completely new calibration samples and the effort required to develop calibration models when changing hardware platforms is minimized. The enabling procedure is based on piecewise direct standardization (PDS) and will be described for the data sets of a quantitative and a qualitative application case study. For this purpose the spectra measured on the FT-NIR laboratory spectrometers were used as "master" data and transferred to the "target" format of the handheld instrument. The quantitative test study refers to transmission spectra of three-component liquid solvent mixtures whereas the qualitative application example encompasses diffuse reflection spectra of six different current polymers. To prove the performance of the transfer procedure for quantitative applications, partial least squares (PLS-1) calibrations were developed for the individual components of the solvent mixtures with spectra transferred from the master to the target instrument and the cross-validation parameters were compared with the corresponding parameters obtained for spectra measured on the master and target instruments, respectively. To test the retention of the discrimination ability of the transferred polymer spectra sets principal component analyses (PCAs) were applied exemplarily for three of the six investigated polymers and their identification was demonstrated by

  12. Photometric calibrations for 21st century science

    Energy Technology Data Exchange (ETDEWEB)

    Kent, Stephen; /Fermilab; Kaiser, Mary Elizabeth; /Johns Hopkins U.; Deustua, Susana E.; /Baltimore, Space Telescope Sci.; Smith, J.Allyn; /Austin Peay State U.; Adelman, Saul; /Citadel Military Coll.; Allam, Sahar S.; /Fermilab; Baptista, Brian; /Indiana U.; Bohlin, Ralph C.; /Baltimore, Space Telescope Sci.; Clem, James L.; /Louisiana State U.; Conley, Alex; /Colorado U.; Edelstein, Jerry; /UC, Berkeley, Space Sci. Dept. /NOAO, Tucson

    2009-02-01

    The answers to fundamental science questions in astrophysics, ranging from the history of the expansion of the universe to the sizes of nearby stars, hinge on our ability to make precise measurements of diverse astronomical objects. As our knowledge of the underlying physics of objects improves along with advances in detectors and instrumentation, the limits on our capability to extract science from measurements is set, not by our lack of understanding of the nature of these objects, but rather by the most mundane of all issues: the precision with which we can calibrate observations in physical units. In principle, photometric calibration is a solved problem - laboratory reference standards such as blackbody furnaces achieve precisions well in excess of those needed for astrophysics. In practice, however, transferring the calibration from these laboratory standards to astronomical objects of interest is far from trivial - the transfer must reach outside the atmosphere, extend over 4{pi} steradians of sky, cover a wide range of wavelengths, and span an enormous dynamic range in intensity. Virtually all spectrophotometric observations today are calibrated against one or more stellar reference sources, such as Vega, which are themselves tied back to laboratory standards in a variety of ways. This system's accuracy is not uniform. Selected regions of the electromagnetic spectrum are calibrated extremely well, but discontinuities of a few percent still exist, e.g., between the optical and infrared. Independently, model stellar atmospheres are used to calibrate the spectra of selected white dwarf stars, e.g. the HST system, but the ultimate accuracy of this system should be verified against laboratory sources. Our traditional standard star systems, while sufficient until now, need to be improved and extended in order to serve future astrophysics experiments. This white paper calls for a program to improve upon and expand the current networks of

  13. Calibration Systems Final Report

    Energy Technology Data Exchange (ETDEWEB)

    Myers, Tanya L.; Broocks, Bryan T.; Phillips, Mark C.

    2006-02-01

    The Calibration Systems project at Pacific Northwest National Laboratory (PNNL) is aimed towards developing and demonstrating compact Quantum Cascade (QC) laser-based calibration systems for infrared imaging systems. These on-board systems will improve the calibration technology for passive sensors, which enable stand-off detection for the proliferation or use of weapons of mass destruction, by replacing on-board blackbodies with QC laser-based systems. This alternative technology can minimize the impact on instrument size and weight while improving the quality of instruments for a variety of missions. The potential of replacing flight blackbodies is made feasible by the high output, stability, and repeatability of the QC laser spectral radiance.

  14. Ibis ground calibration

    Energy Technology Data Exchange (ETDEWEB)

    Bird, A.J.; Barlow, E.J.; Tikkanen, T. [Southampton Univ., School of Physics and Astronomy (United Kingdom); Bazzano, A.; Del Santo, M.; Ubertini, P. [Istituto di Astrofisica Spaziale e Fisica Cosmica - IASF/CNR, Roma (Italy); Blondel, C.; Laurent, P.; Lebrun, F. [CEA Saclay - Sap, 91 - Gif sur Yvette (France); Di Cocco, G.; Malaguti, E. [Istituto di Astrofisica Spaziale e Fisica-Bologna - IASF/CNR (Italy); Gabriele, M.; La Rosa, G.; Segreto, A. [Istituto di Astrofisica Spaziale e Fisica- IASF/CNR, Palermo (Italy); Quadrini, E. [Istituto di Astrofisica Spaziale e Fisica-Cosmica, EASF/CNR, Milano (Italy); Volkmer, R. [Institut fur Astronomie und Astrophysik, Tubingen (Germany)

    2003-11-01

    We present an overview of results obtained from IBIS ground calibrations. The spectral and spatial characteristics of the detector planes and surrounding passive materials have been determined through a series of calibration campaigns. Measurements of pixel gain, energy resolution, detection uniformity, efficiency and imaging capability are presented. The key results obtained from the ground calibration have been: - optimization of the instrument tunable parameters, - determination of energy linearity for all detection modes, - determination of energy resolution as a function of energy through the range 20 keV - 3 MeV, - demonstration of imaging capability in each mode, - measurement of intrinsic detector non-uniformity and understanding of the effects of passive materials surrounding the detector plane, and - discovery (and closure) of various leakage paths through the passive shielding system.

  15. TARGETLESS CAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    L. Barazzetti

    2012-09-01

    Full Text Available In photogrammetry a camera is considered calibrated if its interior orientation parameters are known. These encompass the principal distance, the principal point position and some Additional Parameters used to model possible systematic errors. The current state of the art for automated camera calibration relies on the use of coded targets to accurately determine the image correspondences. This paper presents a new methodology for the efficient and rigorous photogrammetric calibration of digital cameras which does not require any longer the use of targets. A set of images depicting a scene with a good texture are sufficient for the extraction of natural corresponding image points. These are automatically matched with feature-based approaches and robust estimation techniques. The successive photogrammetric bundle adjustment retrieves the unknown camera parameters and their theoretical accuracies. Examples, considerations and comparisons with real data and different case studies are illustrated to show the potentialities of the proposed methodology.

  16. Calibration and validation of a model describing complete autotrophic nitrogen removal in a granular SBR system

    DEFF Research Database (Denmark)

    Vangsgaard, Anna Katrine; Mutlu, Ayten Gizem; Gernaey, Krist

    2013-01-01

    screening of the parameter space proposed by Sin et al. (2008) - to find the best fit of the model to dynamic data. Finally, the calibrated model was validated with an independent data set. CONCLUSION: The presented calibration procedure is the first customized procedure for this type of system...... and is expected to contribute to achieve a fast and effective model calibration, an important enabling tool for various biochemical engineering design, control and operation problems....

  17. Two Methods for Self Calibration of Digital Camera

    Science.gov (United States)

    Sampath, A.; Moe, D.; Christopherson, J.

    2012-07-01

    Photogrammetric mapping using Commercial of the Shelf (COTS) cameras is becoming more popular. Their popularity is augmented by the increasing use of Unmanned Aerial Vehicles (UAV) as a platform for mapping. The mapping precision of these methods can be increased by using a calibrated camera. The USGS/EROS has developed an inexpensive, easy to use method, particularly for calibrating short focal length cameras. The method builds on a self-calibration procedure developed for the USGS EROS Data Center by Pictometry (and augmented by Dr. C.S Fraser), that uses a series of coded targets. These coded targets form different patterns that are imaged from nine different locations with differing camera orientations. A free network solution using collinearity equations is used to determine the calibration parameters. For the smaller focal length COTS cameras, the USGS has developed a procedure that uses a small prototype box that contains these coded targets. The design of the box is discussed, along with best practices for calibration procedure. Results of calibration parameters obtained using the box are compared with the parameters obtained using more established standard procedures.

  18. Iterative Magnetometer Calibration

    Science.gov (United States)

    Sedlak, Joseph

    2006-01-01

    This paper presents an iterative method for three-axis magnetometer (TAM) calibration that makes use of three existing utilities recently incorporated into the attitude ground support system used at NASA's Goddard Space Flight Center. The method combines attitude-independent and attitude-dependent calibration algorithms with a new spinning spacecraft Kalman filter to solve for biases, scale factors, nonorthogonal corrections to the alignment, and the orthogonal sensor alignment. The method is particularly well-suited to spin-stabilized spacecraft, but may also be useful for three-axis stabilized missions given sufficient data to provide observability.

  19. Absolute Radiometric Calibration of ALS Intensity Data: Effects on Accuracy and Target Classification

    Directory of Open Access Journals (Sweden)

    Anssi Krooks

    2011-11-01

    Full Text Available Radiometric calibration of airborne laser scanning (ALS intensity data aims at retrieving a value related to the target scattering properties, which is independent on the instrument or flight parameters. The aim of a calibration procedure is also to be able to compare results from different flights and instruments, but practical applications are sparsely available, and the performance of calibration methods for this purpose needs to be further assessed. We have studied the radiometric calibration with data from three separate flights and two different instruments using external calibration targets. We find that the intensity data from different flights and instruments can be compared to each other only after a radiometric calibration process using separate calibration targets carefully selected for each flight. The calibration is also necessary for target classification purposes, such as separating vegetation from sand using intensity data from different flights. The classification results are meaningful only for calibrated intensity data.

  20. NASA Glenn Icing Research Tunnel: 2014 Cloud Calibration

    Science.gov (United States)

    VanZante, Judith Foss; Ide, Robert F.; Steen, Laura; Acosta, Waldo J.

    2014-01-01

    The results of the December 2013 to February 2014 Icing Research Tunnel full icing cloud calibration are being presented to the SAE AC-9C committee, as represented in the 2014 cloud calibration report. The calibration steps included establishing a uniform cloud and conducting drop size and liquid water content calibrations. The goal of the calibration was to develop a uniform cloud, and to generate a transfer function from the inputs of air speed, spray bar atomizing air pressure and water pressure to the outputs of median volumetric drop diameter and liquid water content. This was done for both 14 CFR Parts 25 and 29, Appendix C (typical icing) and soon-to-be released Appendix O (supercooled large drop) conditions.

  1. Extensions to modeling aerobic carbon degradation using combined respirometric-titrimetric measurements in view of activated sludge model calibration.

    Science.gov (United States)

    Sin, Gürkan; Vanrolleghem, Peter A

    2007-08-01

    Recently a model was introduced to interpret the respirometric (OUR) -titrimetric (Hp) data obtained from aerobic oxidation of different carbon sources in view of calibration of Activated Sludge Model No.1 (ASM1). The model requires, among others, the carbon dioxide transfer rate (CTR) to be relatively constant during aerobic experiments. As CTR is an inherently nonlinear process, this assumption may not hold for certain experimental conditions. Hence, we extended the model to describe the nonlinear CTR behavior. A simple calibration procedure of the CO2 model was developed only using titrimetric data. The identifiable parameter subset of this model when using titrimetric data only contained the first equilibrium constant of the CO2 dissociation, pK1, the initial aqueous CO2 concentration, C(Tinit) and the nitrogen content of biomass, i(NBM). The extended model was then successfully applied to interpret typical data obtained from respirometric-titrimetric measurements with a nonlinear CO2 stripping process. The parameter estimation results using titrimetric data were consistent with the results estimated using respirometric data (OUR) alone or combined OUR and Hp data, thereby supporting the validity of the dynamic CO2 model and its calibration approach. The increased range of applicability and accurate utilization of the titrimetric data are expected to contribute particularly to the improvement of calibration of ASM models using batch experiments.

  2. Effective Calibration of Low-Cost Soil Water Content Sensors

    Directory of Open Access Journals (Sweden)

    Heye Reemt Bogena

    2017-01-01

    Full Text Available Soil water content is a key variable for understanding and modelling ecohydrological processes. Low-cost electromagnetic sensors are increasingly being used to characterize the spatio-temporal dynamics of soil water content, despite the reduced accuracy of such sensors as compared to reference electromagnetic soil water content sensing methods such as time domain reflectometry. Here, we present an effective calibration method to improve the measurement accuracy of low-cost soil water content sensors taking the recently developed SMT100 sensor (Truebner GmbH, Neustadt, Germany as an example. We calibrated the sensor output of more than 700 SMT100 sensors to permittivity using a standard procedure based on five reference media with a known apparent dielectric permittivity (1 < Ka < 34.8. Our results showed that a sensor-specific calibration improved the accuracy of the calibration compared to single “universal” calibration. The associated additional effort in calibrating each sensor individually is relaxed by a dedicated calibration setup that enables the calibration of large numbers of sensors in limited time while minimizing errors in the calibration process.

  3. Smart Calibration of Excavators

    DEFF Research Database (Denmark)

    Bro, Marie; Døring, Kasper; Ellekilde, Lars-Peter

    2005-01-01

    Excavators dig holes. But where is the bucket? The purpose of this report is to treat four different problems concerning calibrations of position indicators for excavators in operation at concrete construction sites. All four problems are related to the question of how to determine the precise ge...

  4. Calibrating Communication Competencies

    Science.gov (United States)

    Surges Tatum, Donna

    2016-11-01

    The Many-faceted Rasch measurement model is used in the creation of a diagnostic instrument by which communication competencies can be calibrated, the severity of observers/raters can be determined, the ability of speakers measured, and comparisons made between various groups.

  5. Entropic calibration revisited

    Energy Technology Data Exchange (ETDEWEB)

    Brody, Dorje C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)]. E-mail: d.brody@imperial.ac.uk; Buckley, Ian R.C. [Centre for Quantitative Finance, Imperial College, London SW7 2AZ (United Kingdom); Constantinou, Irene C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom); Meister, Bernhard K. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)

    2005-04-11

    The entropic calibration of the risk-neutral density function is effective in recovering the strike dependence of options, but encounters difficulties in determining the relevant greeks. By use of put-call reversal we apply the entropic method to the time reversed economy, which allows us to obtain the spot price dependence of options and the relevant greeks.

  6. Calibration with Absolute Shrinkage

    DEFF Research Database (Denmark)

    Øjelund, Henrik; Madsen, Henrik; Thyregod, Poul

    2001-01-01

    is suggested to cope with the singular design matrix most often seen in chemometric calibration. Furthermore, the proposed algorithm may be generalized to all convex norms like Sigma/beta (j)/(gamma) where gamma greater than or equal to 1, i.e. a method that continuously varies from ridge regression...

  7. LOFAR facet calibration

    CERN Document Server

    van Weeren, R J; Hardcastle, M J; Shimwell, T W; Rafferty, D A; Sabater, J; Heald, G; Sridhar, S S; Dijkema, T J; Brunetti, G; Brüggen, M; Andrade-Santos, F; Ogrean, G A; Röttgering, H J A; Dawson, W A; Forman, W R; de Gasperin, F; Jones, C; Miley, G K; Rudnick, L; Sarazin, C L; Bonafede, A; Best, P N; Bîrzan, L; Cassano, R; Chyży, K T; Croston, J H; Ensslin, T; Ferrari, C; Hoeft, M; Horellou, C; Jarvis, M J; Kraft, R P; Mevius, M; Intema, H T; Murray, S S; Orrú, E; Pizzo, R; Simionescu, A; Stroe, A; van der Tol, S; White, G J

    2016-01-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at $\\sim$ 5arcsec resolu...

  8. Measurement System & Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  9. NVLAP calibration laboratory program

    Energy Technology Data Exchange (ETDEWEB)

    Cigler, J.L.

    1993-12-31

    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  10. White light spectral interferometry as a spectrometer calibration tool.

    Science.gov (United States)

    de la Fuente, Raúl

    2014-01-01

    For this paper, we used a white light interferometer in combination with spectral lamps to perform the wavelength calibration of a dispersive spectrometer. Illuminating the spectrometer with suitable spectral lamps gives the wavelength-pixel number relationship at discrete positions of the spectrometer detector array, and the wavelength-dependent phase difference at the output of the white light interferometer allows for a complete spectral calibration at any point on the detector (i.e., for every wavelength in the spectral range of the spectrometer). The details of this new calibration procedure are discussed, and two practical examples exhibiting the robustness of the method are presented. In addition, certain issues relating to minimizing the number of spectral lines used in the calibration procedure are examined.

  11. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini

    2007-03-31

    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  12. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve...... the statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited...

  13. Study on routines and procedures of calibration of clinical dosemeter on gamma radiation beams at the IPEN-CNEN/SP, Brazil; Estudo das rotinas e procedimentos de calibracao de dosimetros clinicos em feixes de radiacao gama no IPEN-CNEN/SP

    Energy Technology Data Exchange (ETDEWEB)

    Damatto, Willian B.; Potiens, Maria da Penha A.; Santos, Gelson P.; Vivolo, Vitor, E-mail: wbdamatto@ipen.b, E-mail: mppalbu@ipen.b, E-mail: gpsantos@ipen.b, E-mail: vivolo@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2009-07-01

    The calibration consists in the comparison of the equipment to be calibrated with other equipment already calibrated by another standard, which means the equipment that will accomplish shall be considered secondary or tertiary, depending on the calibration that will be accomplished. As the clinical dosemeters are used for quality control services in the hospitals, these dosemeters need a reliable measurement precision. Therefore, this work intends to demonstrate the importance of the clinical dosemeter and the calibration, requiring special care face to all clinical set

  14. CP-50 calibration facility radiological safety assessment document

    Energy Technology Data Exchange (ETDEWEB)

    Chilton, M.W.; Hill, R.L.; Eubank, B.F.

    1980-03-01

    The CP-50 Calibration Facility Radiological Safety Assessment document, prepared at the request of the Nevada Operations Office of the US Department of Energy to satisfy provisions of ERDA Manual Chapter 0531, presents design features, systems controls, and procedures used in the operation of the calibration facility. Site and facility characteristics and routine and non-routine operations, including hypothetical incidents or accidents are discussed and design factors, source control systems, and radiation monitoring considerations are described.

  15. 15 CFR 200.108 - Request procedure.

    Science.gov (United States)

    2010-01-01

    ..., PROCEDURES, AND FEES § 200.108 Request procedure. (a) A formal purchase order for the calibration or test should be sent before or at the time the instrument or standard is shipped. The purchase order should... by purchase order or by letter or document authorizing the cost of the work to be billed to...

  16. Finite element model calibration using frequency responses with damping equalization

    Science.gov (United States)

    Abrahamsson, T. J. S.; Kammer, D. C.

    2015-10-01

    Model calibration is a cornerstone of the finite element verification and validation procedure, in which the credibility of the model is substantiated by positive comparison with test data. The calibration problem, in which the minimum deviation between finite element model data and experimental data is searched for, is normally characterized as being a large scale optimization problem with many model parameters to solve for and with deviation metrics that are nonlinear in these parameters. The calibrated parameters need to be found by iterative procedures, starting from initial estimates. Sometimes these procedures get trapped in local deviation function minima and do not converge to the globally optimal calibration solution that is searched for. The reason for such traps is often the multi-modality of the problem which causes eigenmode crossover problems in the iterative variation of parameter settings. This work presents a calibration formulation which gives a smooth deviation metric with a large radius of convergence to the global minimum. A damping equalization method is suggested to avoid the mode correlation and mode pairing problems that need to be solved in many other model updating procedures. By this method, the modal damping of a test data model and the finite element model is set to be the same fraction of critical modal damping. Mode pairing for mapping of experimentally found damping to the finite element model is thus not needed. The method is combined with model reduction for efficiency and employs the Levenberg-Marquardt minimizer with randomized starts to achieve the calibration solution. The performance of the calibration procedure, including a study of parameter bias and variance under noisy data conditions, is demonstrated by two numerical examples.

  17. Marine X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren Liedtke; Rasmussen, Michael R.

    2012-01-01

    Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis, and real time control purposes. In these contexts, it is allimportant that the radar data is well calibrated and adjusted in order to obtain valid quantitative precipitation...... estimates. This paper presents some of the challenges in small marine X-band radar calibration by comparing three calibration procedures for assessing the relationship between radar and rain gauge data. Validation shows similar results for precipitation volumes but more diverse results on peak rain...

  18. Calibration of a physically based, spatially distributed hydrological model in a glacierized basin: On the use of knowledge from glaciometeorological processes to constrain model parameters

    Science.gov (United States)

    Ragettli, S.; Pellicciotti, F.

    2012-03-01

    In the Dry Andes of central Chile, summer water resources originate mostly from snowmelt and ice melt. We use the physically based, spatially distributed hydrological model TOPKAPI to study the exchange between glaciers and climate in the upper Aconcagua River Basin during the summer season and identify the model parameters that are robust and transferable and those that are more dependent on calibration. TOPKAPI has recently been adapted to incorporate an enhanced temperature index approach for snow and ice melting. We suggest a calibration procedure that allows calibration of parameters in three steps by separating parameters governing distinct processes. We evaluate the parameters' transferability in time and in space by applying the model at two spatial scales. TOPKAPI's ability to simulate the relevant processes is tested against meteorological, ablation, and glacier runoff data measured on Juncal Norte Glacier during two glacier ablation seasons. The model was applied successfully to the climatic setting of the Dry Andes once its parameters were recalibrated. We found a clear distinction between parameters that are stable in time and those that need recalibration. The parameters of the melt model are transferable from one season to the other, while the parameters governing the extrapolation of meteorological input data and the routing of glacier meltwater need recalibration from one season to the other. Sensitivity analysis revealed that the model is most sensitive to the temperature lapse rate governing the extrapolation of air temperature from point measurements to the glacier scale and to the melt parameter that multiplies the shortwave radiation balance.

  19. Two-Step Camera Calibration Method Developed for Micro UAV'S

    Science.gov (United States)

    Gašparović, M.; Gajski, D.

    2016-06-01

    The development of unmanned aerial vehicles (UAVs) and continuous price reduction of unmanned systems attracted us to this research. Professional measuring systems are dozens of times more expensive and often heavier than "amateur", non-metric UAVs. For this reason, we tested the DJI Phantom 2 Vision Plus UAV. Phantom's smaller mass and velocity can develop less kinetic energy in relation to the professional measurement platforms, which makes it potentially less dangerous for use in populated areas. In this research, we wanted to investigate the ability of such non-metric UAV and find the procedures under which this kind of UAV may be used for the photogrammetric survey. It is important to emphasize that UAV is equipped with an ultra wide-angle camera with 14MP sensor. Calibration of such cameras is a complex process. In the research, a new two-step process is presented and developed, and the results are compared with standard one-step camera calibration procedure. Two-step process involves initially removed distortion on all images, and then uses these images in the phototriangulation with self-calibration. The paper presents statistical indicators which proved that the proposed two-step process is better and more accurate procedure for calibrating those types of cameras than standard one-step calibration. Also, we suggest two-step calibration process as the standard for ultra-wideangle cameras for unmanned aircraft.

  20. Whole-Body Vibration Sensor Calibration Using a Six-Degree of Freedom Robot

    Directory of Open Access Journals (Sweden)

    Sarah Cation

    2011-01-01

    Full Text Available Exposure to whole-body vibration (WBV is associated with a wide variety of health disorders and as a result WBV levels are frequently assessed. Literature outlining WBV accelerations rarely address the calibration techniques and procedures used for WBV sensors to any depth, nor are any detailed information provided regarding such procedures or sensor calibration ranges. The purpose of this paper is to describe a calibration method for a 6 DOF transducer using a hexapod robot. Also described is a separate motion capture technique used to verify the calibration for acceleration values obtained which were outside the robot calibration range in order to include an acceptable calibration range for WBV environments. The sensor calibrated in this study used linear (Y=mX calibration equations resulting in r2 values greater than 0.97 for maximum and minimum acceleration amplitudes of up to ±8 m/s2 and maximum and minimum velocity amplitudes up to ±100°/s. The motion capture technique verified that the translational calibrations held for accelerations up to ±4 g. Thus, the calibration procedures were shown to calibrate the sensor through the expected range for 6-DOF WBV field measurements for off-road vehicles even when subjected to shocks as a result of high speed travel over rough terrain.

  1. The Calibration Reference Data System

    Science.gov (United States)

    Greenfield, P.; Miller, T.

    2016-07-01

    We describe a software architecture and implementation for using rules to determine which calibration files are appropriate for calibrating a given observation. This new system, the Calibration Reference Data System (CRDS), replaces what had been previously used for the Hubble Space Telescope (HST) calibration pipelines, the Calibration Database System (CDBS). CRDS will be used for the James Webb Space Telescope (JWST) calibration pipelines, and is currently being used for HST calibration pipelines. CRDS can be easily generalized for use in similar applications that need a rules-based system for selecting the appropriate item for a given dataset; we give some examples of such generalizations that will likely be used for JWST. The core functionality of the Calibration Reference Data System is available under an Open Source license. CRDS is briefly contrasted with a sampling of other similar systems used at other observatories.

  2. Geometric calibration of the SND detector electromagnetic calorimeter

    CERN Document Server

    Korol, A A

    2016-01-01

    This paper presents the design, implementation and validation of the software alignment procedure used to perform geometric calibration of the electromagnetic calorimeter with respect to the tracking system of the Spherical Neutral Detector (SND) which is used for HEP experiments at the VEPP-2000 $e^{+}e^{-}$ collider (BINP, Novosibirsk). This procedure is based on the mathematical model describing the relative calorimeter position. The parameter values are determined by minimizing a $\\chi^{2}$ function using the difference between directions reconstructed in these two subdetectors for the $e^{+}e^{-}\\rightarrow e^{+}e^{-}$ scattering data events. The results of the calibration and its application to the data and MC simulation fit the model and give an improvement of the reconstructed particle parameters. They are used already in the actual experimental data analysis and more realistic MC simulation. We think that the ideas implemented in the calibration procedure can be useful for other experiments with hete...

  3. Lidar calibration experiments

    DEFF Research Database (Denmark)

    Ejsing Jørgensen, Hans; Mikkelsen, T.; Streicher, J.

    1997-01-01

    A series of atmospheric aerosol diffusion experiments combined with lidar detection was conducted to evaluate and calibrate an existing retrieval algorithm for aerosol backscatter lidar systems. The calibration experiments made use of two (almost) identical mini-lidar systems for aerosol cloud...... detection to test the reproducibility and uncertainty of lidars. Lidar data were obtained from both single-ended and double-ended Lidar configurations. A backstop was introduced in one of the experiments and a new method was developed where information obtained from the backstop can be used in the inversion...... algorithm. Independent in-situ aerosol plume concentrations were obtained from a simultaneous tracer gas experiment with SF6, and comparisons with the two lidars were made. The study shows that the reproducibility of the lidars is within 15%, including measurements from both sides of a plume...

  4. Optical tweezers absolute calibration

    CERN Document Server

    Dutra, R S; Neto, P A Maia; Nussenzveig, H M

    2014-01-01

    Optical tweezers are highly versatile laser traps for neutral microparticles, with fundamental applications in physics and in single molecule cell biology. Force measurements are performed by converting the stiffness response to displacement of trapped transparent microspheres, employed as force transducers. Usually, calibration is indirect, by comparison with fluid drag forces. This can lead to discrepancies by sizable factors. Progress achieved in a program aiming at absolute calibration, conducted over the past fifteen years, is briefly reviewed. Here we overcome its last major obstacle, a theoretical overestimation of the peak stiffness, within the most employed range for applications, and we perform experimental validation. The discrepancy is traced to the effect of primary aberrations of the optical system, which are now included in the theory. All required experimental parameters are readily accessible. Astigmatism, the dominant effect, is measured by analyzing reflected images of the focused laser spo...

  5. Radiative Transfer Model for Contaminated Rough Surfaces

    Science.gov (United States)

    2013-02-01

    plot of Figure 8 shows three sharp spectral features (in the LWIR region) that were used for calibration . 1000 1500 2000 2500 3000 3500 0 0.1 0.2...transfer, reflectance, rough surface, BRDF, Kramers-Kronig, penetration depth, fill factor, infrared, LWIR , MWIR, absorption coefficient, scattering...and the calibrated α are plotted in red, and green, respectively

  6. Astrid-2 SSC ASUMagnetic Calibration

    DEFF Research Database (Denmark)

    Primdahl, Fritz

    1997-01-01

    Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory.......Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory....

  7. Calibration Facilities for NIF

    Energy Technology Data Exchange (ETDEWEB)

    Perry, T.S.

    2000-06-15

    The calibration facilities will be dynamic and will change to meet the needs of experiments. Small sources, such as the Manson Source should be available to everyone at any time. Carrying out experiments at Omega is providing ample opportunity for practice in pre-shot preparation. Hopefully, the needs that are demonstrated in these experiments will assure the development of (or keep in service) facilities at each of the laboratories that will be essential for in-house preparation for experiments at NIF.

  8. Mesoscale hybrid calibration artifact

    Science.gov (United States)

    Tran, Hy D.; Claudet, Andre A.; Oliver, Andrew D.

    2010-09-07

    A mesoscale calibration artifact, also called a hybrid artifact, suitable for hybrid dimensional measurement and the method for make the artifact. The hybrid artifact has structural characteristics that make it suitable for dimensional measurement in both vision-based systems and touch-probe-based systems. The hybrid artifact employs the intersection of bulk-micromachined planes to fabricate edges that are sharp to the nanometer level and intersecting planes with crystal-lattice-defined angles.

  9. The Calibration Home Base for Imaging Spectrometers

    Directory of Open Access Journals (Sweden)

    Johannes Felix Simon Brachmann

    2016-08-01

    Full Text Available The Calibration Home Base (CHB is an optical laboratory designed for the calibration of imaging spectrometers for the VNIR/SWIR wavelength range. Radiometric, spectral and geometric calibration as well as the characterization of sensor signal dependency on polarization are realized in a precise and highly automated fashion. This allows to carry out a wide range of time consuming measurements in an ecient way. The implementation of ISO 9001 standards in all procedures ensures a traceable quality of results. Spectral measurements in the wavelength range 380–1000 nm are performed to a wavelength uncertainty of +- 0.1 nm, while an uncertainty of +-0.2 nm is reached in the wavelength range 1000 – 2500 nm. Geometric measurements are performed at increments of 1.7 µrad across track and 7.6 µrad along track. Radiometric measurements reach an absolute uncertainty of +-3% (k=1. Sensor artifacts, such as caused by stray light will be characterizable and correctable in the near future. For now, the CHB is suitable for the characterization of pushbroom sensors, spectrometers and cameras. However, it is planned to extend the CHBs capabilities in the near future such that snapshot hyperspectral imagers can be characterized as well. The calibration services of the CHB are open to third party customers from research institutes as well as industry.

  10. Studies Concerning the ATLAS IBL Calibration Architecture

    CERN Document Server

    Kretz, Moritz; Kugel, Andreas

    With the commissioning of the Insertable B-Layer (IBL) in 2013 at the ATLAS experiment 12~million additional pixels will be added to the current Pixel Detector. While the idea of employing pairs of VME based Read-Out Driver (ROD) and Back of Crate (BOC) cards in the read-out chain remains unchanged, modifications regarding the IBL calibration procedure were introduced to overcome current hardware limitations. The analysis of calibration histograms will no longer be performed on the RODs, but on an external computing farm that is connected to the RODs via Ethernet. This thesis contributes to the new IBL calibration procedure and presents a concept for a scalable software and hardware architecture. An embedded system targeted to the ROD FPGAs is realized for sending data from the RODs to the fit farm servers and benchmarks are carried out with a Linux based networking stack, as well as a standalone software stack. Furthermore, the histogram fitting algorithm currently being employed on the Pixel Detector RODs i...

  11. Calibration of Underwater Sound Transducers

    Directory of Open Access Journals (Sweden)

    H.R.S. Sastry

    1983-07-01

    Full Text Available The techniques of calibration of underwater sound transducers for farfield, near-field and closed environment conditions are reviewed in this paper .The design of acoustic calibration tank is mentioned. The facilities available at Naval Physical & Oceanographic Laboratory, Cochin for calibration of transducers are also listed.

  12. Precision Measurement and Calibration. Volume 1. Statistical Concepts and Procedures

    Science.gov (United States)

    1969-02-01

    Sons, New York. N.Y.). Shewhart, Walter A. (19t39), Statistical Method frin the Galilei , Galileo (1638), Discorsi e Dirnostrazioni Matema- Viewpoint... Galilei . Galileo (1898). Discorsi e Dimostraziorii Matemnaticlie Shiewhart, Walter A. (1941), (Contribiution of statistics to Initorno a Dine Ninove Scienze...Le Opere i Galileo Galilei the science of engineering, University of Pennsylvania Bi- (Edizione Nazionale) Vill, pp 39-448, Firenze. centennial

  13. RORASC: Software for the rotated-random-scan calibration procedure

    NARCIS (Netherlands)

    Janssen PHM; CWM

    1995-01-01

    De release-versie 1.0 van het programma RORASC wordt gepresenteerd, tezamen met instructies en richtlijnen voor installatie en gebruik van de software. RORASC speelt een essentiele rol bij de 'rotated-random-scan' methode voor modelkalibratie. De software is geschreven in standaard FOR

  14. A PROCEDURE TO CALIBRATE AND PERFORM THE BENDER ELEMENT TEST

    Directory of Open Access Journals (Sweden)

    JAVIER FERNANDO CAMACHO-TAUTA

    2012-01-01

    Full Text Available El método de ensayo bender element es utilizado en laboratorio para obtener la velocidad de la onda cortante de suelos. A pesar de su aparente simplicidad, aún no existe una norma para este ensayo. Existen factores que pueden tener alguna incidencia sobre el resultado. Estos factores son tenidos en cuenta para diseñar y construir un sistema de ensayo en una cámara triaxial. Se propone un método para calibrar el sistema mediante la utilización de barras de aluminio de diferente longitud y se hacen recomendaciones para la ejecución del ensayo, en lo relacionado con la selección de la frecuencia del pulso sinusoidal utilizado para producir la perturbación. Finalmente, se presentan ensayos efectuados sobre dos muestras de arena.

  15. Method of biodosimeter calibration for orbital flight

    Science.gov (United States)

    Vladislav, Petrov

    A biodosimetry method, based on estimation of an absorbed dose on the basis of assessment of unstable aberration frequency in the lymphocytes of human's peripheral blood is used broadly in ground conditions for analysis of accidental exposure to personal and citizens. A calibration curve giving the relationship between aberration frequency (generally dicentrics and centric rings) and an absorbed dose in blood samples is used for assessment of crewmember exposure. As a rule gamma - rays corresponding to character of exposure in such accidents are used for these goals. At the same time the space radiation fields are formed mainly by charged particles for which the character of effect on the body cells and tissues differs strongly from that of gamma - rays. As biodosimetry is a relative method of dose measurement it is necessary to obtain a calibration curve corresponding to the conditions in which the measurements will be performed. That is a calibration curve for space application should give a relationship between aberration frequency and a dose formed by radiation field equal to that on the spacecraft trajectory. The report contains a method of obtaining a calibration curve for a case of an orbital flight on the ISS trajectory. The radiobiological basis of the method consists of relationships between chromosomal aberration frequency in human blood lymphocytes and an absorbed dose of protons with four energies (50 MeV, 150 MeV, 400 MeV, 625 MeV) obtained in the accelerator's experiments. Due to the fact that we had experimental data only for protons the calibration curve was obtained for the proton component on the ISS orbit which is mainly formed by trapped protons. Dose spectrum for this energy distribution of protons was calculated and weighting coefficients for taking into account the input of dose of protons with various energies in forming total frequency of chromosomal aberrations were obtained on its basis. The procedure of obtaining such weighting

  16. A New Method for Common Calibration of Sun-Sky-Lunar Photometer

    Science.gov (United States)

    Li, Kaitao; Li, Zhengqiang; Li, Donghui; Xu, Hua; Xie, Yisong; Li, Li; Chen, Xingfeng; Ma, Yan

    2016-04-01

    A new calibration method is introduced to transfer extraterrestrial calibration coefficients to the moon measurements for a new sun-sky-lunar photometer, trade name CE318-T from CIMEL. The new transfer method has no relationship with lunar phase, therefore, the precision of the results is improved, and error analysis suggests that the uncertainty of the transferred method is about 2.2-2.6%, smaller than the lunar Langley calibrations. At the same time, the calibration time is also saved. The Sun-Sky-Lunar photometer numbered #1202 and located on the roof of Institute of remote sensing and digital earth (RADI) in Beijing was used in this study. The extraterrestrial calibration coefficients were got by using Langley calibration performed at Ali with a height of 5053 m above sea level in Tibet. The new lunar calibration coefficients were obtained with the new transfer method. And then the nocturnal AODs were calculated, which are well consistent with the daytime observations. The differences between two AODs obtained with transferred calibration coefficients and lunar Langley method were also compared in this paper. In this study, Lidar observation results was also presented to compare with the lunar observations, the results show that the nocturnal AODs have the same variation tendency with the Lidar observations.

  17. In situ ``artificial plasma'' calibration of tokamak magnetic sensors

    Science.gov (United States)

    Shiraki, D.; Levesque, J. P.; Bialek, J.; Byrne, P. J.; DeBono, B. A.; Mauel, M. E.; Maurer, D. A.; Navratil, G. A.; Pedersen, T. S.; Rath, N.

    2013-06-01

    A unique in situ calibration technique has been used to spatially calibrate and characterize the extensive new magnetic diagnostic set and close-fitting conducting wall of the High Beta Tokamak-Extended Pulse (HBT-EP) experiment. A new set of 216 Mirnov coils has recently been installed inside the vacuum chamber of the device for high-resolution measurements of magnetohydrodynamic phenomena including the effects of eddy currents in the nearby conducting wall. The spatial positions of these sensors are calibrated by energizing several large in situ calibration coils in turn, and using measurements of the magnetic fields produced by the various coils to solve for each sensor's position. Since the calibration coils are built near the nominal location of the plasma current centroid, the technique is referred to as an "artificial plasma" calibration. The fitting procedure for the sensor positions is described, and results of the spatial calibration are compared with those based on metrology. The time response of the sensors is compared with the evolution of the artificial plasma current to deduce the eddy current contribution to each signal. This is compared with simulations using the VALEN electromagnetic code, and the modeled copper thickness profiles of the HBT-EP conducting wall are adjusted to better match experimental measurements of the eddy current decay. Finally, the multiple coils of the artificial plasma system are also used to directly calibrate a non-uniformly wound Fourier Rogowski coil on HBT-EP.

  18. Geometric calibration of the circle-plus-arc trajectory.

    Science.gov (United States)

    Hoppe, Stefan; Noo, Frédéric; Dennerlein, Frank; Lauritsch, Günter; Hornegger, Joachim

    2007-12-07

    In this paper, a novel geometric calibration method for C-arm cone-beam scanners is presented which allows the calibration of the circle-plus-arc trajectory. The main idea is the separation of the trajectory into two circular segments (circle segment and arc segment) which are calibrated independently. This separation makes it possible to reuse a calibration phantom which has been successfully applied in clinical environments to calibrate numerous routinely used C-arm systems. For each trajectory segment, the phantom is placed in an optimal position. The two calibration results are then combined by computing the transformation the phantom underwent between the independent calibration runs. This combination can be done in a post-processing step by using standard linear algebra. The method is not limited to circle-plus-arc trajectories and works for any calibration procedure in which the phantom has a preferred orientation with respect to a trajectory segment. Results are presented for both simulated as well as real data acquired with a C-arm system. We also present the first image reconstruction results for the circle-plus-arc trajectory using real C-arm data.

  19. Spectrophotometer spectral bandwidth calibration with absorption bands crystal standard.

    Science.gov (United States)

    Soares, O D; Costa, J L

    1999-04-01

    A procedure for calibration of a spectral bandwidth standard for high-resolution spectrophotometers is described. Symmetrical absorption bands for a crystal standard are adopted. The method relies on spectral band shape fitting followed by a convolution with the slit function of the spectrophotometer. A reference spectrophotometer is used to calibrate the spectral bandwidth standard. Bandwidth calibration curves for a minimum spectral transmission factor relative to the spectral bandwidth of the reference spectrophotometer are derived for the absorption bands at the wavelength of the band absorption maximum. The family of these calibration curves characterizes the spectral bandwidth standard. We calibrate the spectral bandwidth of a spectrophotometer with respect to the reference spectrophotometer by determining the spectral transmission factor minimum at every calibrated absorption band of the bandwidth standard for the nominal instrument values of the spectral bandwidth. With reference to the standard spectral bandwidth calibration curves, the relation of the spectral bandwidth to the reference spectrophotometer is determined. We determine the discrepancy in the spectrophotometers' spectral bandwidths by averaging the spectral bandwidth discrepancies relative to the standard calibrated values found at the absorption bands considered. A weighted average of the uncertainties is taken.

  20. Summary of the steps involved in the calibration of a Spinner anemometer

    DEFF Research Database (Denmark)

    Demurtas, Giorgio; Friis Pedersen, Troels

    The present report is a practical guide that summarizes the steps involved in the calibration of a spinner anemometer. For each step is brie y recalled the objective, which document describes the procedure to use, and what actions must follow the calibration.......The present report is a practical guide that summarizes the steps involved in the calibration of a spinner anemometer. For each step is brie y recalled the objective, which document describes the procedure to use, and what actions must follow the calibration....

  1. Towards automatic calibration of 2-dimensional flood propagation models

    Directory of Open Access Journals (Sweden)

    P. Fabio

    2009-11-01

    Full Text Available Hydraulic models for flood propagation description are an essential tool in many fields, e.g. civil engineering, flood hazard and risk assessments, evaluation of flood control measures, etc. Nowadays there are many models of different complexity regarding the mathematical foundation and spatial dimensions available, and most of them are comparatively easy to operate due to sophisticated tools for model setup and control. However, the calibration of these models is still underdeveloped in contrast to other models like e.g. hydrological models or models used in ecosystem analysis. This has basically two reasons: first, the lack of relevant data against the models can be calibrated, because flood events are very rarely monitored due to the disturbances inflicted by them and the lack of appropriate measuring equipment in place. Secondly, especially the two-dimensional models are computationally very demanding and therefore the use of available sophisticated automatic calibration procedures is restricted in many cases. This study takes a well documented flood event in August 2002 at the Mulde River in Germany as an example and investigates the most appropriate calibration strategy for a full 2-D hyperbolic finite element model. The model independent optimiser PEST, that gives the possibility of automatic calibrations, is used. The application of the parallel version of the optimiser to the model and calibration data showed that a it is possible to use automatic calibration in combination of 2-D hydraulic model, and b equifinality of model parameterisation can also be caused by a too large number of degrees of freedom in the calibration data in contrast to a too simple model setup. In order to improve model calibration and reduce equifinality a method was developed to identify calibration data with likely errors that obstruct model calibration.

  2. Recent developments in radon metrology: new aspects in the calibration of radon, thoron and progeny devices.

    Science.gov (United States)

    Röttger, A; Honig, A

    2011-05-01

    Due to the importance of reliable measurements of radon activity concentration, one of the past developments in metrology was applied to the field of radon, thus meeting two basic needs: (1) the harmonisation of metrology within the scope of the mutual recognition arrangement, an arrangement drawn up by the International Committee of Weights and Measures for the mutual recognition of national standards and of calibrations issued by national metrology institutes and (2) the increased demands of the European Atomic Energy Community (EURATOM) directive, transferred into national radiation protection regulations with regard to natural radioactivity and its quality-assured measurements. This paper gives an overview of typical technical procedures in the radon-measuring technique group of PTB, covering all aspects of reference atmospheres (primary standards) for radon, thoron and their respective progenies.

  3. MAGNETIC GRADIOMETRY: Instrumentation, Calibration and Applications

    DEFF Research Database (Denmark)

    Merayo, Jose Maria Garcia

    The description of the single axis magnetic gradiometer based on two Compensation Detector Coil (CDC) fluxgate ringcore sensors separated 20cm introduces the subject of magnetic gradiometry. Despite its good properties and high precision of less than 1nT, the calibration procedures...... are not straightforward. Firstly, the monoaxial character does not provide the vectorial information on the magnetic field. Secondly, one of the sensors measures the ambient magnetic field and this is used to compensate the field in both sensors. Several methods have been developed for its characterization...... and the calibration of the gradient measurement is achieved by the use of a magnetic dipole pattern of strength 2mAm2. In a coil facility, the gradient can be determined with an RMS value of 0.3nT/m.The ultra high sensitivity magnetic triaxial gradiometer has been constructed by employing another approach. Two...

  4. Viscometry for liquids calibration of viscometers

    CERN Document Server

    Gupta, S V

    2014-01-01

    This book is written for scientists involved in the calibration of viscometers. A detailed description for stepping up procedures to establish the viscosity scale and obtaining sets of master viscometers is given in the book. Uncertainty considerations for standard oils of known viscosity are presented. The modern viscometers based on principles of  tuning fork, ultrasonic, PZT, plate waves, Love waves, micro-cantilever and vibration of optical fiber are discussed to inspire the reader to further research and to generate improved versions. The primary standard for viscosity is pure water. Measurements of its viscosity with accuracy/uncertainty achieved are described. The principles of rotational and oscillation viscometers are explained to enhance the knowledge in calibration work. Devices used for specific materials and viscosity in non SI units are discussed with respect to the need to correlate viscosity values obtained by various devices. The description of commercial viscometers meets the needs of the u...

  5. Determination of calibration function in thermal field flow fractionation under thermal field programming.

    Science.gov (United States)

    Pastil, Luisa; Ventosa, Edgar A; Mingozzi, Ines; Dondi, Francesco

    2006-05-01

    A new procedure for determining the calibration function able to relate retention and operative parameters to molecular weight of the species in thermal field flow (ThFFF) under thermal field programming (TFP) conditions is presented. The procedure involves determining the average values of retention parameters under TFP and determining a numerical function related to the temperature variations that occur during TFP. The calibration parameters are obtained by a procedure fitting the retention and operative parameters that hold true at the beginning of the TFP. The procedure is closely related to the one previously developed to calibrate the retention time axis under TFP ThFFF and, together, they constitute a full calibration procedure. Experimental validation was performed with reference to polystyrene (PS)-decalin and PS-THF systems. The calibration functions here obtained were compared to those derived by the classical procedure at constant thermal field ThFFF to obtain the calibration function at variable cold wall temperatures. Excellent agreement was found in all cases proving "universality" of the ThFFF calibration concept, i.e. it is independent of the particular system on which it was determined and can thus be extended to ThFFF operating under TFP. The new procedure is simpler than the classical one since it requires less precision in setting the instrumentation and can be obtained with fewer experiments. The potential applications for the method are discussed.

  6. Application of Orthogonal Signal Correction for Near Infrared Spectroscopic Calibration Transfer of Crude Protein of Straw Silage%基于正交信号校正的秸秆青贮饲料粗蛋白近红外分析模型传递方法

    Institute of Scientific and Technical Information of China (English)

    刘贤; 韩鲁佳; 杨增玲; 黄光群

    2012-01-01

    探讨了基于不同数据预处理方法的正交信号校正在秸杆饲料近红外光谱模型传递中的应用.以141个秸杆青贮饲料样品为研究对象,以其粗蛋白含量为目标参数,研究了基于无处理、局部中心化、全局中心化和Z-score标准化预处理方法的正交信号校正,在源仪器(SPECTRUM ONE NTS)和目标仪器1(ANTA-RIS)与目标仪器2(FOSS 6500)之间的模型传递效果.实验表明:对于两台傅里叶变换型近红外光谱仪,采用局部中心化、全局中心化和Z-score标准化预处理方法的正交信号校正均可成功实现模型传递,其中局部中心化和全局中心化法的作用效果基本一致,且优于Z-score标准化法.对于傅立叶变换和光栅型近红外光谱仪,全局中心化的作用效果明显优于其它3组处理效果,且只有全局中心化预处理的正交信号校正传递后的模型可用于实际预测.%This study is focus on the orthogonal signal correction based on different pretreatment methods for near infrared spectroscopy (NIRS) calibration transfer of straw silage. Up to 141 straw silage samples with known crude protein content and two Fourier transform instruments (master instrument: SPECTRUM ONE NTS. Slave instrument 1: ANTARIS), one scanning grating instrument (slave instrument 2: FOSS 6500) were involved. Four pretreatment methods, control, local centering, global centering and Z-score standardization, for orthogonal signal correction were tested and evaluated. The results showed that the orthogonal signal correction with the local centering, the global centering and the Z-score standardization methods worked successful for calibration transfer between two Fourier transform instruments. The local centering and the global centering gave almost equally good results which was better than that of Z-score standardization method. However, for calibration transfer between Fourier transform and scanning grating instruments, only global centering

  7. Secondary calibration laboratory for ionizing radiation laboratory accreitation program National Institute of Standards and Technology National Voluntary Laboratory Accreditation Program

    Energy Technology Data Exchange (ETDEWEB)

    Martin, P.R.

    1993-12-31

    This paper presents an overview of the procedures and requirements for accreditation under the Secondary Calibration Laboratory for Ionizing Radiation Program (SCLIR LAP). The requirements for a quality system, proficiency testing and the onsite assessment are discussed. The purpose of the accreditation program is to establish a network of secondary calibration laboratories that can provide calibrations traceable to the primary national standards.

  8. CALIBRATED HYDRODYNAMIC MODEL

    Directory of Open Access Journals (Sweden)

    Sezar Gülbaz

    2015-01-01

    Full Text Available The land development and increase in urbanization in a watershed affect water quantityand water quality. On one hand, urbanization provokes the adjustment of geomorphicstructure of the streams, ultimately raises peak flow rate which causes flood; on theother hand, it diminishes water quality which results in an increase in Total SuspendedSolid (TSS. Consequently, sediment accumulation in downstream of urban areas isobserved which is not preferred for longer life of dams. In order to overcome thesediment accumulation problem in dams, the amount of TSS in streams and inwatersheds should be taken under control. Low Impact Development (LID is a BestManagement Practice (BMP which may be used for this purpose. It is a land planningand engineering design method which is applied in managing storm water runoff inorder to reduce flooding as well as simultaneously improve water quality. LID includestechniques to predict suspended solid loads in surface runoff generated over imperviousurban surfaces. In this study, the impact of LID-BMPs on surface runoff and TSS isinvestigated by employing a calibrated hydrodynamic model for Sazlidere Watershedwhich is located in Istanbul, Turkey. For this purpose, a calibrated hydrodynamicmodel was developed by using Environmental Protection Agency Storm WaterManagement Model (EPA SWMM. For model calibration and validation, we set up arain gauge and a flow meter into the field and obtain rainfall and flow rate data. Andthen, we select several LID types such as retention basins, vegetative swales andpermeable pavement and we obtain their influence on peak flow rate and pollutantbuildup and washoff for TSS. Consequently, we observe the possible effects ofLID on surface runoff and TSS in Sazlidere Watershed.

  9. ALTEA: The instrument calibration

    Energy Technology Data Exchange (ETDEWEB)

    Zaconte, V. [INFN and University of Rome Tor Vergata, Department of Physics, Via della Ricerca Scientifica 1, 00133 Rome (Italy)], E-mail: livio.narici@roma2.infn.it; Belli, F.; Bidoli, V.; Casolino, M.; Di Fino, L.; Narici, L.; Picozza, P.; Rinaldi, A. [INFN and University of Rome Tor Vergata, Department of Physics, Via della Ricerca Scientifica 1, 00133 Rome (Italy); Sannita, W.G. [DISM, University of Genova, Genova (Italy); Department of Psychiatry, SUNY, Stoony Brook, NY (United States); Finetti, N.; Nurzia, G.; Rantucci, E.; Scrimaglio, R.; Segreto, E. [Department of Physics, University and INFN, L' Aquila (Italy); Schardt, D. [GSI/Biophysik, Darmstadt (Germany)

    2008-05-15

    The ALTEA program is an international and multi-disciplinary project aimed at studying particle radiation in space environment and its effects on astronauts' brain functions, as the anomalous perception of light flashes first reported during Apollo missions. The ALTEA space facility includes a 6-silicon telescopes particle detector, and is onboard the International Space Station (ISS) since July 2006. In this paper, the detector calibration at the heavy-ion synchrotron SIS18 at GSI Darmstadt will be presented and compared to the Geant 3 Monte Carlo simulation. Finally, the results of a neural network analysis that was used for ion discrimination on fragmentation data will also be presented.

  10. Dynamic Torque Calibration Unit

    Science.gov (United States)

    Agronin, Michael L.; Marchetto, Carl A.

    1989-01-01

    Proposed dynamic torque calibration unit (DTCU) measures torque in rotary actuator components such as motors, bearings, gear trains, and flex couplings. Unique because designed specifically for testing components under low rates. Measures torque in device under test during controlled steady rotation or oscillation. Rotor oriented vertically, supported by upper angular-contact bearing and lower radial-contact bearing that floats axially to prevent thermal expansion from loading bearings. High-load capacity air bearing available to replace ball bearings when higher load capacity or reduction in rate noise required.

  11. Cross calibration of IRS-P4 OCM satellite sensor

    Digital Repository Service at National Institute of Oceanography (India)

    Suresh, T.; Desa, E.; Mascarenhas, A.A.M.Q.; Matondkar, S.G.P.; Naik, P.; Nayak, S.R.

    The cross calibration of ocean color satellite sensor, IRS-P4 OCM using the radiative transfer code, with SeaWiFS as a reference are presented here. Since the bands of IRS-P4 OCM are identical to those of SeaWiFS and SeaWiFS has been continuously...

  12. Calibration of the BEV GPS Receiver by Using TWSTFT

    Science.gov (United States)

    2008-12-01

    B. Blanzano, and O. Koudelka, 2008, “Time transfer with nanosecond accuracy for the realization of International Atomic Time,” Metrologia , 45, 185...TAI computation,” Metrologia , 45, 35-45 [4] A. Niessner, 14 January 2008, “BEV Report to BIPM of the calibration result for the GPS receivers

  13. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  14. Summary of KOMPSAT-5 Calibration and Validation

    Science.gov (United States)

    Yang, D.; Jeong, H.; Lee, S.; Kim, B.

    2013-12-01

    including pointing, relative and absolute calibration as well as geolocation accuracy determination. The absolute calibration will be accomplished by determining absolute radiometric accuracy using already deployed trihedral corner reflectors on calibration and validation sites located southeast from Ulaanbaatar, Mongolia. To establish a measure for the assess the final image products, geolocation accuracies of image products with different imaging modes will be determined by using deployed point targets and available Digital Terrain Model (DTM), and on different image processing levels. In summary, this paper will present calibration and validation activities performed during the LEOP and IOT of KOMPSAT-5. The methodology and procedure of calibration and validation will be explained as well as its results. Based on the results, the applications of SAR image products on geophysical processes will be also discussed.

  15. A Simple Accelerometer Calibrator

    Science.gov (United States)

    Salam, R. A.; Islamy, M. R. F.; Munir, M. M.; Latief, H.; Irsyam, M.; Khairurrijal

    2016-08-01

    High possibility of earthquake could lead to the high number of victims caused by it. It also can cause other hazards such as tsunami, landslide, etc. In that case it requires a system that can examine the earthquake occurrence. Some possible system to detect earthquake is by creating a vibration sensor system using accelerometer. However, the output of the system is usually put in the form of acceleration data. Therefore, a calibrator system for accelerometer to sense the vibration is needed. In this study, a simple accelerometer calibrator has been developed using 12 V DC motor, optocoupler, Liquid Crystal Display (LCD) and AVR 328 microcontroller as controller system. The system uses the Pulse Wave Modulation (PWM) form microcontroller to control the motor rotational speed as response to vibration frequency. The frequency of vibration was read by optocoupler and then those data was used as feedback to the system. The results show that the systems could control the rotational speed and the vibration frequencies in accordance with the defined PWM.

  16. Hot-wire calibration in subsonic/transonic flow regimes

    Science.gov (United States)

    Nagabushana, K. A.; Ash, Robert L.

    1995-01-01

    A different approach for calibrating hot-wires, which simplifies the calibration procedure and reduces the tunnel run-time by an order of magnitude was sought. In general, it is accepted that the directly measurable quantities in any flow are velocity, density, and total temperature. Very few facilities have the capability of varying the total temperature over an adequate range. However, if the overheat temperature parameter, a(sub w), is used to calibrate the hot-wire then the directly measurable quantity, voltage, will be a function of the flow variables and the overheat parameter i.e., E = f(u,p,a(sub w), T(sub w)) where a(sub w) will contain the needed total temperature information. In this report, various methods of evaluating sensitivities with different dependent and independent variables to calibrate a 3-Wire hot-wire probe using a constant temperature anemometer (CTA) in subsonic/transonic flow regimes is presented. The advantage of using a(sub w) as the independent variable instead of total temperature, t(sub o), or overheat temperature parameter, tau, is that while running a calibration test it is not necessary to know the recovery factor, the coefficients in a wire resistance to temperature relationship for a given probe. It was deduced that the method employing the relationship E = f (u,p,a(sub w)) should result in the most accurate calibration of hot wire probes. Any other method would require additional measurements. Also this method will allow calibration and determination of accurate temperature fluctuation information even in atmospheric wind tunnels where there is no ability to obtain any temperature sensitivity information at present. This technique greatly simplifies the calibration process for hot-wires, provides the required calibration information needed in obtaining temperature fluctuations, and reduces both the tunnel run-time and the test matrix required to calibrate hotwires. Some of the results using the above techniques are presented

  17. Resent Progress in Research on Calibration Instrument for Radioactive Aerosol Monitor

    Institute of Scientific and Technical Information of China (English)

    CHEN; Xi-lin; CHEN; Yong-yong; WU; Chang-ping; XING; Yu; MENG; Jun; YANG; Qiao-ling

    2013-01-01

    Radioactive aerosol monitors are widely used in monitoring the radioactivity concentration of the artificial nuclides in gaseous effluents from the nuclear facilities.An on-developing calibration instrument for radioactive aerosol monitors consists of an α and β aerosol generating unit,aerosol transferring unit,measurement unit of radioactivity concentration of aerosol for instruments calibrated and the waste gas

  18. Leveraging microwave polarization information for calibration of a land data assimilation system

    Science.gov (United States)

    This letter contributes a new approach to calibrating a tau-omega radiative transfer model coupled to a land surface model with low frequency (< 10 GHz) microwave brightness temperature (TB) observations. The problem of calibrating this system is generally poorlyposed because various parameter combi...

  19. Calibration of high-heat-flux sensors in a solar furnace

    Energy Technology Data Exchange (ETDEWEB)

    Ballestrin, J.; Rodriguez-Alonso, M.; Rodriguez, J.; Canadas, I. [Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas (CIEMAT), Almeria (Spain); Barbero, F.J. [Almeria Univ., Dept. de Fisica (Spain); Langley, L.W.; Barnes, A. [Vatell Corporation, Christiansburg, VA (United States)

    2006-12-15

    The most common sensors used for the measurement of high solar irradiance are the Gardon gauges, which are usually calibrated using a black body at a certain temperature as the radiant source. This calibration procedure is assumed to produce a systematic error when solar irradiance measurements are taken using these sensors. This paper demonstrates a calorimetric method for calibrating these high-heat-flux gauges in a solar furnace. This procedure has enabled these sensors to be calibrated under concentrated solar radiation at higher irradiances under non-laboratory conditions in the CIEMAT solar furnace at the Plataforma Solar de Almeria. Working at higher irradiances has allowed the uncertainty in the calibration constant of these sensors to be reduced. This work experimentally confirms the predicted systematic errors committed when measuring high solar irradiances using Gardon sensors calibrated with a black body. (authors)

  20. Calibration of high-heat-flux sensors in a solar furnace

    Science.gov (United States)

    Ballestrín, J.; Rodríguez-Alonso, M.; Rodríguez, J.; Cañadas, I.; Barbero, F. J.; Langley, L. W.; Barnes, A.

    2006-12-01

    The most common sensors used for the measurement of high solar irradiance are the Gardon gauges, which are usually calibrated using a black body at a certain temperature as the radiant source. This calibration procedure is assumed to produce a systematic error when solar irradiance measurements are taken using these sensors. This paper demonstrates a calorimetric method for calibrating these high-heat-flux gauges in a solar furnace. This procedure has enabled these sensors to be calibrated under concentrated solar radiation at higher irradiances under non-laboratory conditions in the CIEMAT solar furnace at the Plataforma Solar de Almería. Working at higher irradiances has allowed the uncertainty in the calibration constant of these sensors to be reduced. This work experimentally confirms the predicted systematic errors committed when measuring high solar irradiances using Gardon sensors calibrated with a black body.

  1. Calibration of Short Range 2D Laser Range Finder for 3D SLAM Usage

    Directory of Open Access Journals (Sweden)

    Petr Olivka

    2016-01-01

    Full Text Available The laser range finder is one of the most essential sensors in the field of robotics. The laser range finder provides an accurate range measurement with high angular resolution. However, the short range scanners require an additional calibration to achieve the abovementioned accuracy. The calibration procedure described in this work provides an estimation of the internal parameters of the laser range finder without requiring any special three-dimensional targets. This work presents the use of a short range URG-04LX scanner for mapping purposes and describes its calibration. The precision of the calibration was checked in an environment with known ground truth values and the results were statistically evaluated. The benefits of the calibration are also demonstrated in the practical applications involving the segmentation of the environment. The proposed calibration method is complex and detects all major manufacturing inaccuracies. The procedure is suitable for easy integration into the current manufacturing process.

  2. Measurement quality assurance for beta particle calibrations at NIST

    Energy Technology Data Exchange (ETDEWEB)

    Soares, C.G.; Pruitt, J.S. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    Standardized beta-particle fields have been established in an international standard and have been adopted for use in several U.S. dosimeter and instrument testing standards. Calibration methods and measurement quality assurance procedures employed at the National Institute of Standards and Technology (NIST) for beta-particle calibrations in these reference fields are discussed. The calibration facility including the NIST-automated extrapolation ionization chamber is described, and some sample results of calibrations are shown. Methods for establishing and maintaining traceability to NIST of secondary laboratories are discussed. Currently, there are problems in finding a good method for routine testing of traceability to NIST. Some examples of past testing methods are given and solutions to this problem are proposed.

  3. Exploring uncertainty in the calibration of the molecular clock.

    Science.gov (United States)

    Warnock, Rachel C M; Yang, Ziheng; Donoghue, Philip C J

    2012-02-23

    Calibration is a critical step in every molecular clock analysis but it has been the least considered. Bayesian approaches to divergence time estimation make it possible to incorporate the uncertainty in the degree to which fossil evidence approximates the true time of divergence. We explored the impact of different approaches in expressing this relationship, using arthropod phylogeny as an example for which we established novel calibrations. We demonstrate that the parameters distinguishing calibration densities have a major impact upon the prior and posterior of the divergence times, and it is critically important that users evaluate the joint prior distribution of divergence times used by their dating programmes. We illustrate a procedure for deriving calibration densities in Bayesian divergence dating through the use of soft maximum constraints.

  4. Magnetic information calibrates celestial cues during migration.

    Science.gov (United States)

    Sandberg; Bäckman; Moore; Lõhmus

    2000-10-01

    Migratory birds use celestial and geomagnetic directional information to orient on their way between breeding and wintering areas. Cue-conflict experiments involving these two orientation cue systems have shown that directional information can be transferred from one system to the other by calibration. We designed experiments with four species of North American songbirds to: (1) examine whether these species calibrate orientation information from one system to the other; and (2) determine whether there are species-specific differences in calibration. Migratory orientation was recorded with two different techniques, cage tests and free-flight release tests, during autumn migration. Cage tests at dusk in the local geomagnetic field revealed species-specific differences: red-eyed vireo, Vireo olivaceus, and northern waterthrush, Seiurus noveboracensis, selected seasonally appropriate southerly directions whereas indigo bunting, Passerina cyanea, and grey catbird, Dumetella carolinensis, oriented towards the sunset direction. When tested in deflected magnetic fields, vireos and waterthrushes responded by shifting their orientation according to the deflection of the magnetic field, but buntings and catbirds failed to show any response to the treatment. In release tests, all four species showed that they had recalibrated their star compass on the basis of the magnetic field they had just experienced in the cage tests. Since release tests were done in the local geomagnetic field it seems clear that once the migratory direction is determined, most likely during the twilight period, the birds use their recalibrated star compass for orientation at departure. Copyright 2000 The Association for the Study of Animal Behaviour.

  5. Systematic Calibration for a Backpacked Spherical Photogrammetry Imaging System

    Science.gov (United States)

    Rau, J. Y.; Su, B. W.; Hsiao, K. W.; Jhan, J. P.

    2016-06-01

    A spherical camera can observe the environment for almost 720 degrees' field of view in one shoot, which is useful for augmented reality, environment documentation, or mobile mapping applications. This paper aims to develop a spherical photogrammetry imaging system for the purpose of 3D measurement through a backpacked mobile mapping system (MMS). The used equipment contains a Ladybug-5 spherical camera, a tactical grade positioning and orientation system (POS), i.e. SPAN-CPT, and an odometer, etc. This research aims to directly apply photogrammetric space intersection technique for 3D mapping from a spherical image stereo-pair. For this purpose, several systematic calibration procedures are required, including lens distortion calibration, relative orientation calibration, boresight calibration for direct georeferencing, and spherical image calibration. The lens distortion is serious on the ladybug-5 camera's original 6 images. Meanwhile, for spherical image mosaicking from these original 6 images, we propose the use of their relative orientation and correct their lens distortion at the same time. However, the constructed spherical image still contains systematic error, which will reduce the 3D measurement accuracy. Later for direct georeferencing purpose, we need to establish a ground control field for boresight/lever-arm calibration. Then, we can apply the calibrated parameters to obtain the exterior orientation parameters (EOPs) of all spherical images. In the end, the 3D positioning accuracy after space intersection will be evaluated, including EOPs obtained by structure from motion method.

  6. Accurate and simple calibration of DLP projector systems

    Science.gov (United States)

    Wilm, Jakob; Olesen, Oline V.; Larsen, Rasmus

    2014-03-01

    Much work has been devoted to the calibration of optical cameras, and accurate and simple methods are now available which require only a small number of calibration targets. The problem of obtaining these parameters for light projectors has not been studied as extensively and most current methods require a camera and involve feature extraction from a known projected pattern. In this work we present a novel calibration technique for DLP Projector systems based on phase shifting profilometry projection onto a printed calibration target. In contrast to most current methods, the one presented here does not rely on an initial camera calibration, and so does not carry over the error into projector calibration. A radial interpolation scheme is used to convert features coordinates into projector space, thereby allowing for a very accurate procedure. This allows for highly accurate determination of parameters including lens distortion. Our implementation acquires printed planar calibration scenes in less than 1s. This makes our method both fast and convenient. We evaluate our method in terms of reprojection errors and structured light image reconstruction quality.

  7. Absolute calibration of a multilayer-based XUV diagnostic

    CERN Document Server

    Stuik, R; Tümmler, J; Bijkerk, F

    2002-01-01

    A portable, universal narrowband XUV diagnostic suitable for calibration of various XUV light sources, was built, tested and fully calibrated. The diagnostic allows measurement of the absolute XUV energy and average power in two selected wavelength bands, at 11.4 and 13.4 nm. In addition, the pulse-to-pulse and long-term XUV stability of the source can be assessed, as well as the contamination of multilayer XUV optics exposed to the source. This paper describes the full calibration procedure: all optical elements were calibrated at the wavelength of operation by Physikalisch-Technische Bundesanstalt at the storage ring Bessy II, a full analysis of geometrical factors was done, and the influence of the spectral emissivity of the source on the calibration was analyzed in detail. The calibration was performed both for the centroid wavelength as for the full bandwidth of the diagnostic. The total uncertainty in the absolute calibration allowed measurement of source characteristics with an uncertainty of less than...

  8. Modeling of germanium detector and its sourceless calibration

    Directory of Open Access Journals (Sweden)

    Steljić Milijana

    2008-01-01

    Full Text Available The paper describes the procedure of adapting a coaxial high-precision germanium detector to a device with numerical calibration. The procedure includes the determination of detector dimensions and establishing the corresponding model of the system. In order to achieve a successful calibration of the system without the usage of standard sources, Monte Carlo simulations were performed to determine its efficiency and pulse-height response function. A detailed Monte Carlo model was developed using the MCNP-5.0 code. The obtained results have indicated that this method represents a valuable tool for the quantitative uncertainty analysis of radiation spectrometers and gamma-ray detector calibration, thus minimizing the need for the deployment of radioactive sources.

  9. Self-calibrating lateral scanning white-light interferometer

    Energy Technology Data Exchange (ETDEWEB)

    Munteanu, Florin

    2010-04-20

    Lateral scanning white-light interferometry represents an attractive alternative to the standard white-light interferometry. Its main advantage over the latter procedure consists in the ability to scan large samples continuously, without the need of a cumbersome stitching procedure. Presently, the main drawback in the path of large-scale industrial acceptance of this method is the need for careful calibration of the tilt angle prior to each measurement. A novel self-calibration approach is presented. Using the data acquired during the normal scanning process, the need of an initial tilt angle calibration is eliminated and on-the-fly system adjustments for the best signal-to-noise ratio can be performed without an increase in the measurement time dictated by recalibration.

  10. Conditional Versus Unconditional Procedures for Sample-Free Item Analysis

    Science.gov (United States)

    Wright, Benjamin D.; Douglas, Graham A.

    1977-01-01

    Two procedures for Rasch, sample-free item calibration are reviewed and compared for accuracy. The theoretically ideal "conditional" procedure is impractical for more than fifteen items. The more practical but biased "unconditional" procedure is discussed in detail. (Author/JKS)

  11. Internal Water Vapor Photoacoustic Calibration

    Science.gov (United States)

    Pilgrim, Jeffrey S.

    2009-01-01

    Water vapor absorption is ubiquitous in the infrared wavelength range where photoacoustic trace gas detectors operate. This technique allows for discontinuous wavelength tuning by temperature-jumping a laser diode from one range to another within a time span suitable for photoacoustic calibration. The use of an internal calibration eliminates the need for external calibrated reference gases. Commercial applications include an improvement of photoacoustic spectrometers in all fields of use.

  12. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Kristensen, L.; Jensen, G.; Hansen, A.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statisticalsignificance of the calibration expressions....... It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometersmay take more than one month in order to have wind speeds covering a sufficiently large magnitude range...

  13. Radiological Calibration and Standards Facility

    Data.gov (United States)

    Federal Laboratory Consortium — PNNL maintains a state-of-the-art Radiological Calibration and Standards Laboratory on the Hanford Site at Richland, Washington. Laboratory staff provide expertise...

  14. SURF Model Calibration Strategy

    Energy Technology Data Exchange (ETDEWEB)

    Menikoff, Ralph [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-03-10

    SURF and SURFplus are high explosive reactive burn models for shock initiation and propagation of detonation waves. They are engineering models motivated by the ignition & growth concept of high spots and for SURFplus a second slow reaction for the energy release from carbon clustering. A key feature of the SURF model is that there is a partial decoupling between model parameters and detonation properties. This enables reduced sets of independent parameters to be calibrated sequentially for the initiation and propagation regimes. Here we focus on a methodology for tting the initiation parameters to Pop plot data based on 1-D simulations to compute a numerical Pop plot. In addition, the strategy for tting the remaining parameters for the propagation regime and failure diameter is discussed.

  15. RX130 Robot Calibration

    Science.gov (United States)

    Fugal, Mario

    2012-10-01

    In order to create precision magnets for an experiment at Oak Ridge National Laboratory, a new reverse engineering method has been proposed that uses the magnetic scalar potential to solve for the currents necessary to produce the desired field. To make the magnet it is proposed to use a copper coated G10 form, upon which a drill, mounted on a robotic arm, will carve wires. The accuracy required in the manufacturing of the wires exceeds nominal robot capabilities. However, due to the rigidity as well as the precision servo motor and harmonic gear drivers, there are robots capable of meeting this requirement with proper calibration. Improving the accuracy of an RX130 to be within 35 microns (the accuracy necessary of the wires) is the goal of this project. Using feedback from a displacement sensor, or camera and inverse kinematics it is possible to achieve this accuracy.

  16. Stability study of standards used for calibration of the spectrophotometer wavelength scale

    Science.gov (United States)

    Debossan, L. F.; Carvalho, E. M. S.; Souza, M. A.; Gomes, J. F. S.

    2016-07-01

    The calibration of spectrophotometers is a procedure recommended by international standards to provide quality assurance of results and traceability. Due to its intrinsic properties, holmium oxide filters are indicated as reference standards for calibrating the wavelength scale of such equipment. This paper presents a study aiming to assess the repeatability and drift of holmium oxide standard filters calibrated in the Radiometry and Photometry Laboratory (Laraf) of Inmetro in order to verify their stability.

  17. 40 CFR 89.320 - Carbon monoxide analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... range. If the deviation exceeds these limits, the best-fit non-linear equation which represents the data... interference, system check, and calibration test procedures specified in 40 CFR part 1065 may be used in lieu... is 2 percent or less of the value at each non-zero data point and within ± 0.3 percent of full...

  18. Construction of calibration pads facility, Walker Field, Grand Junction, Colorado

    Energy Technology Data Exchange (ETDEWEB)

    Ward, D.L.

    1978-08-01

    A gamma-ray spectrometer facility was completed at Walker Field Airport, Grand Junction, Colorado, in November 1976. This report describes spectrometers and their calibration, the construction of the spectrometer facility, the radioelement concentrations, procedures for using the facilites, and environmental considerations. (LK)

  19. 40 CFR 86.221-94 - Hydrocarbon analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Hydrocarbon analyzer calibration. 86.221-94 Section 86.221-94 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... New Medium-Duty Passenger Vehicles; Cold Temperature Test Procedures § 86.221-94 Hydrocarbon...

  20. 40 CFR 86.224-94 - Carbon dioxide analyzer calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Carbon dioxide analyzer calibration. 86.224-94 Section 86.224-94 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... New Medium-Duty Passenger Vehicles; Cold Temperature Test Procedures § 86.224-94 Carbon...

  1. Absolute Calibration and Characterization of the Multiband Imaging Photometer for Spitzer. I. The Stellar Calibrator Sample and the 24 micron Calibration

    CERN Document Server

    Engelbracht, C W; Su, K Y L; Rho, J; Rieke, G H; Muzerolle, J; Padgett, D L; Hines, D C; Gordon, K D; Fadda, D; Noriega-Crespo, A; Kelly, D M; Latter, W B; Hinz, J L; Misselt, K A; Morrison, J E; Stansberry, J A; Shupe, D L; Stolovy, S; Wheaton, Wm A; Young, E T; Neugebauer, G; Wachter, S; Pérez-González, P G; Frayer, D T; Marleau, F R

    2007-01-01

    We present the stellar calibrator sample and the conversion from instrumental to physical units for the 24 micron channel of the Multiband Imaging Photometer for Spitzer (MIPS). The primary calibrators are A stars, and the calibration factor based on those stars is 4.54*10^{-2} MJy sr^{-1} (DN/s)^{-1}, with a nominal uncertainty of 2%. We discuss the data-reduction procedures required to attain this accuracy; without these procdures, the calibration factor obtained using the automated pipeline at the Spitzer Science Center is 1.6% +/- 0.6% lower. We extend this work to predict 24 micron flux densities for a sample of 238 stars which covers a larger range of flux densities and spectral types. We present a total of 348 measurements of 141 stars at 24 micron. This sample covers a factor of ~460 in 24 micron flux density, from 8.6 mJy up to 4.0 Jy. We show that the calibration is linear over that range with respect to target flux and background level. The calibration is based on observations made using 3-second e...

  2. Calibration of Nanopositioning Stages

    Directory of Open Access Journals (Sweden)

    Ning Tan

    2015-12-01

    Full Text Available Accuracy is one of the most important criteria for the performance evaluation of micro- and nanorobots or systems. Nanopositioning stages are used to achieve the high positioning resolution and accuracy for a wide and growing scope of applications. However, their positioning accuracy and repeatability are not well known and difficult to guarantee, which induces many drawbacks for many applications. For example, in the mechanical characterisation of biological samples, it is difficult to perform several cycles in a repeatable way so as not to induce negative influences on the study. It also prevents one from controlling accurately a tool with respect to a sample without adding additional sensors for closed loop control. This paper aims at quantifying the positioning repeatability and accuracy based on the ISO 9283:1998 standard, and analyzing factors influencing positioning accuracy onto a case study of 1-DoF (Degree-of-Freedom nanopositioning stage. The influence of thermal drift is notably quantified. Performances improvement of the nanopositioning stage are then investigated through robot calibration (i.e., open-loop approach. Two models (static and adaptive models are proposed to compensate for both geometric errors and thermal drift. Validation experiments are conducted over a long period (several days showing that the accuracy of the stage is improved from typical micrometer range to 400 nm using the static model and even down to 100 nm using the adaptive model. In addition, we extend the 1-DoF calibration to multi-DoF with a case study of a 2-DoF nanopositioning robot. Results demonstrate that the model efficiently improved the 2D accuracy from 1400 nm to 200 nm.

  3. Dental Procedures.

    Science.gov (United States)

    Ramponi, Denise R

    2016-01-01

    Dental problems are a common complaint in emergency departments in the United States. There are a wide variety of dental issues addressed in emergency department visits such as dental caries, loose teeth, dental trauma, gingival infections, and dry socket syndrome. Review of the most common dental blocks and dental procedures will allow the practitioner the opportunity to make the patient more comfortable and reduce the amount of analgesia the patient will need upon discharge. Familiarity with the dental equipment, tooth, and mouth anatomy will help prepare the practitioner for to perform these dental procedures.

  4. A reference material for dynamic displacement calibration

    Science.gov (United States)

    Davighi, A.; Hack, E.; Patterson, E.; Whelan, M.

    2010-06-01

    Calibration of displacement and strain measurement systems is an essential step in providing traceability and confidence in stress and strain distributions obtained from experiment and used to validate simulations employed in engineering design. Reference materials provide a simple, well-defined distribution of the measured quantity that can be traced to an international standard and can be used to assess the uncertainty associated with the measurement system. Previous work has established a reference material and procedure for calibrating optical systems for measuring static, in-plane strain distributions and also demonstrated its use. A new effort is in progress to extend this work to the measurement of three-dimensional displacement distributions induced by cyclic and dynamic loading, including transients and large-scale deformation. The first step in this effort has been to define both the essential and desirable attributes of a reference material for calibrating systems capable of measurements of dynamic displacement and strain. An international consortium of research laboratories, system designers, manufacturers and end-users has identified a list of attributes and members of the experimental mechanics community have been asked to weight the importance of these attributes. The attributes are being utilised to evaluate candidate designs for the reference material which have been generated through a series of brain-storming sessions within the consortium.

  5. Absolute Radiometric Calibration of KOMPSAT-3A

    Science.gov (United States)

    Ahn, H. Y.; Shin, D. Y.; Kim, J. S.; Seo, D. C.; Choi, C. U.

    2016-06-01

    This paper presents a vicarious radiometric calibration of the Korea Multi-Purpose Satellite-3A (KOMPSAT-3A) performed by the Korea Aerospace Research Institute (KARI) and the Pukyong National University Remote Sensing Group (PKNU RSG) in 2015.The primary stages of this study are summarized as follows: (1) A field campaign to determine radiometric calibrated target fields was undertaken in Mongolia and South Korea. Surface reflectance data obtained in the campaign were input to a radiative transfer code that predicted at-sensor radiance. Through this process, equations and parameters were derived for the KOMPSAT-3A sensor to enable the conversion of calibrated DN to physical units, such as at-sensor radiance or TOA reflectance. (2) To validate the absolute calibration coefficients for the KOMPSAT-3A sensor, we performed a radiometric validation with a comparison of KOMPSAT-3A and Landsat-8 TOA reflectance using one of the six PICS (Libya 4). Correlations between top-of-atmosphere (TOA) radiances and the spectral band responses of the KOMPSAT-3A sensors at the Zuunmod, Mongolia and Goheung, South Korea sites were significant for multispectral bands. The average difference in TOA reflectance between KOMPSAT-3A and Landsat-8 image over the Libya 4, Libya site in the red-green-blue (RGB) region was under 3%, whereas in the NIR band, the TOA reflectance of KOMPSAT-3A was lower than the that of Landsat-8 due to the difference in the band passes of two sensors. The KOMPSAT-3Aensor includes a band pass near 940 nm that can be strongly absorbed by water vapor and therefore displayed low reflectance. Toovercome this, we need to undertake a detailed analysis using rescale methods, such as the spectral bandwidth adjustment factor.

  6. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-07-01

    Accurate solar radiation data sets are critical to reducing the expenses associated with mitigating performance risk for solar energy conversion systems, and they help utility planners and grid system operators understand the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of calibration methodologies and the resulting calibration responsivities provided by radiometric calibration service providers such as the National Renewable Energy Laboratory (NREL) and manufacturers of radiometers. Some of these radiometers are calibrated indoors, and some are calibrated outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The reference radiometer calibrations are traceable to the World Radiometric Reference. These different methods of calibration demonstrated 1% to 2% differences in solar irradiance measurement. Analyzing these values will ultimately assist in determining the uncertainties of the radiometer data and will assist in developing consensus on a standard for calibration.

  7. A critical comparison of systematic calibration protocols for activated sludge models: a SWOT analysis.

    Science.gov (United States)

    Sin, Gürkan; Van Hulle, Stijn W H; De Pauw, Dirk J W; van Griensven, Ann; Vanrolleghem, Peter A

    2005-07-01

    Modelling activated sludge systems has gained an increasing momentum after the introduction of activated sludge models (ASMs) in 1987. Application of dynamic models for full-scale systems requires essentially a calibration of the chosen ASM to the case under study. Numerous full-scale model applications have been performed so far which were mostly based on ad hoc approaches and expert knowledge. Further, each modelling study has followed a different calibration approach: e.g. different influent wastewater characterization methods, different kinetic parameter estimation methods, different selection of parameters to be calibrated, different priorities within the calibration steps, etc. In short, there was no standard approach in performing the calibration study, which makes it difficult, if not impossible, to (1) compare different calibrations of ASMs with each other and (2) perform internal quality checks for each calibration study. To address these concerns, systematic calibration protocols have recently been proposed to bring guidance to the modeling of activated sludge systems and in particular to the calibration of full-scale models. In this contribution four existing calibration approaches (BIOMATH, HSG, STOWA and WERF) will be critically discussed using a SWOT (Strengths, Weaknesses, Opportunities, Threats) analysis. It will also be assessed in what way these approaches can be further developed in view of further improving the quality of ASM calibration. In this respect, the potential of automating some steps of the calibration procedure by use of mathematical algorithms is highlighted.

  8. Calibration and modelling of the SODART-OXS Bragg spectrometer onboard the SRG satellite

    DEFF Research Database (Denmark)

    Halm, Ingolf; Wiebicke, Hans-Joachim; Christensen, Finn Erland

    1998-01-01

    -5 keV, 0.5-1.2 keV, and 0.16-0.42 keV. All types of crystals besides Si being an ideal crystal (with parameters which were calculated) have been calibrated individually and after gluing onto the Bragg panel. The X-ray calibration procedures and results are discussed below. A ray-tracing program using...

  9. Combined calibration and sensitivity analysis for a water quality model of the Biebrza River, Poland

    NARCIS (Netherlands)

    Perk, van der M.; Bierkens, M.F.P.

    1995-01-01

    A study was performed to quantify the error in results of a water quality model of the Biebrza River, Poland, due to uncertainties in calibrated model parameters. The procedure used in this study combines calibration and sensitivity analysis. Finally,the model was validated to test the model capabil

  10. Calibration of well-type ionization chambers; Calibracao de camaras de ionizacao do tipo poco

    Energy Technology Data Exchange (ETDEWEB)

    Alves, C.F.E.; Leite, S.P.; Pires, E.J.; Magalhaes, L.A.G.; David, M.G.; Almeida, C.E. de, E-mail: cfealves@gmail.com [Universidade do Estado do Rio de Janeiro (UERJ), Rio de Janeiro, RJ (Brazil). Lab. de Ciencias Radiologicas; Di Prinzio, R. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

    2015-07-01

    This paper presents the methodology developed by the Laboratorio de Ciencias Radiologicas and presently in use for determining of the calibration coefficient for well-type chambers used in the dosimetry of {sup 192}Ir high dose rate sources. Uncertainty analysis involving the calibration procedure are discussed. (author)

  11. Use and Calibration of Capacitive RH Sensors for the Hygrometric Control of the CMS Tracker

    CERN Document Server

    Fossa, Marco

    2003-01-01

    definition of rules to be identified in order to perform an overall calibration procedure devoted to the selection of proper instrumentation able to assure long period operations under well-established measurement uncertainty.

  12. A Rapid Calibration Technique for Scanning Line-Structured Laser Sensor

    Institute of Scientific and Technical Information of China (English)

    Li Tao; Changku Sun; Zhiqin Xu; Wei Wei

    2003-01-01

    A novel procedure to calibrate the scanning line-structured laser sensor is presented. A drone composed of two orthogonal planes is designed, with the result that camera parameters and light-plane equation parameters is achieved simultaneously.

  13. Radiometrie recalibration procedure for landsat-5 thematic mapper data

    Science.gov (United States)

    Chander, G.; Micijevic, E.; Hayes, R.W.; Barsi, J.A.

    2008-01-01

    The Landsat-5 (L5) satellite was launched on March 01, 1984, with a design life of three years. Incredibly, the L5 Thematic Mapper (TM) has collected data for 23 years. Over this time, the detectors have aged, and its radiometric characteristics have changed since launch. The calibration procedures and parameters have also changed with time. Revised radiometric calibrations have improved the radiometric accuracy of recently processed data; however, users with data that were processed prior to the calibration update do not benefit from the revisions. A procedure has been developed to give users the ability to recalibrate their existing Level 1 (L1) products without having to purchase reprocessed data from the U.S. Geological Survey (USGS). The accuracy of the recalibration is dependent on the knowledge of the prior calibration applied to the data. The ""Work Order" file, included with standard National Land Archive Production System (NLAFS) data products, gives parameters that define the applied calibration. These are the Internal Calibrator (IC) calibration parameters or the default prelaunch calibration, if there were problems with the IC calibration. This paper details the recalibration procedure for data processed using IC, in which users have the Work Order file. ?? 2001 IEEE.

  14. Calibration of the Super-Kamiokande Detector

    CERN Document Server

    Abe, K; Iida, T; Iyogi, K; Kameda, J; Kishimoto, Y; Koshio, Y; Marti, Ll; Miura, M; Moriyama, S; Nakahata, M; Nakano, Y; Nakayama, S; Obayashi, Y; Sekiya, H; Shiozawa, M; Suzuki, Y; Takeda, A; Takenaga, Y; Tanaka, H; Tomura, T; Ueno, K; Wendell, R A; Yokozawa, T; Irvine, T J; Kaji, H; Kajita, T; Kaneyuki, K; Lee, K P; Nishimura, Y; Okumura, K; McLachlan, T; Labarga, L; Kearns, E; Raaf, J L; Stone, J L; Sulak, L R; Berkman, S; Tanaka, H A; Tobayama, S; Goldhaber, M; Bays, K; Carminati, G; Kropp, W R; Mine, S; Renshaw, A; Smy, M B; Sobel, H W; Ganezer, K S; Hill, J; Keig, W E; Jang, J S; Kim, J Y; Lim, I T; Hong, N; Akiri, T; Albert, J B; Himmel, A; Scholberg, K; Walter, C W; Wongjirad, T; Ishizuka, T; Tasaka, S; Learned, J G; Matsuno, S; Smith, S N; Hasegawa, T; Ishida, T; Ishii, T; Kobayashi, T; Nakadaira, T; Nakamura, K; Nishikawa, K; Oyama, Y; Sakashita, K; Sekiguchi, T; Tsukamoto, T; Suzuki, A T; Takeuchi, Y; Huang, K; Ieki, K; Ikeda, M; Kikawa, T; Kubo, H; Minamino, A; Murakami, A; Nakaya, T; Otani, M; Suzuki, K; Takahashi, S; Fukuda, Y; Choi, K; Itow, Y; Mitsuka, G; Miyake, M; Mijakowski, P; Tacik, R; Hignight, J; Imber, J; Jung, C K; Taylor, I; Yanagisawa, C; Idehara, Y; Ishino, H; Kibayashi, A; Mori, T; Sakuda, M; Yamaguchi, R; Yano, T; Kuno, Y; Kim, S B; Yang, B S; Okazawa, H; Choi, Y; Nishijima, K; Koshiba, M; Totsuka, Y; Yokoyama, M; Martens, K; Vagins, M R; Martin, J F; de Perio, P; Konaka, A; Wilking, M J; Chen, S; Heng, Y; Sui, H; Yang, Z; Zhang, H; Zhenwei, Y; Connolly, K; Dziomba, M; Wilkes, R J

    2013-01-01

    Procedures and results on hardware level detector calibration in Super-Kamiokande (SK) are presented in this paper. In particular, we report improvements made in our calibration methods for the experimental phase IV in which new readout electronics have been operating since 2008. The topics are separated into two parts. The first part describes the determination of constants needed to interpret the digitized output of our electronics so that we can obtain physical numbers such as photon counts and their arrival times for each photomultiplier tube (PMT). In this context, we developed an in-situ procedure to determine high-voltage settings for PMTs in large detectors like SK, as well as a new method for measuring PMT quantum efficiency and gain in such a detector. The second part describes the modeling of the detector in our Monte Carlo simulation, including in particular the optical properties of its water target and their variability over time. Detailed studies on the water quality are also presented. As a re...

  15. Novel calibration algorithm for a three-axis strapdown magnetometer.

    Science.gov (United States)

    Liu, Yan Xia; Li, Xi Sheng; Zhang, Xiao Juan; Feng, Yi Bo

    2014-05-14

    A complete error calibration model with 12 independent parameters is established by analyzing the three-axis magnetometer error mechanism. The said model conforms to an ellipsoid restriction, the parameters of the ellipsoid equation are estimated, and the ellipsoid coefficient matrix is derived. However, the calibration matrix cannot be determined completely, as there are fewer ellipsoid parameters than calibration model parameters. Mathematically, the calibration matrix derived from the ellipsoid coefficient matrix by a different matrix decomposition method is not unique, and there exists an unknown rotation matrix R between them. This paper puts forward a constant intersection angle method (angles between the geomagnetic field and gravitational field are fixed) to estimate R. The Tikhonov method is adopted to solve the problem that rounding errors or other errors may seriously affect the calculation results of R when the condition number of the matrix is very large. The geomagnetic field vector and heading error are further corrected by R. The constant intersection angle method is convenient and practical, as it is free from any additional calibration procedure or coordinate transformation. In addition, the simulation experiment indicates that the heading error declines from ±1° calibrated by classical ellipsoid fitting to ±0.2° calibrated by a constant intersection angle method, and the signal-to-noise ratio is 50 dB. The actual experiment exhibits that the heading error is further corrected from ±0.8° calibrated by the classical ellipsoid fitting to ±0.3° calibrated by a constant intersection angle method.

  16. Efficiency transfer using the GEANT4 code of CERN for HPGe gamma spectrometry.

    Science.gov (United States)

    Chagren, S; Ben Tekaya, M; Reguigui, N; Gharbi, F

    2016-01-01

    In this work we apply the GEANT4 code of CERN to calculate the peak efficiency in High Pure Germanium (HPGe) gamma spectrometry using three different procedures. The first is a direct calculation. The second corresponds to the usual case of efficiency transfer between two different configurations at constant emission energy assuming a reference point detection configuration and the third, a new procedure, consists on the transfer of the peak efficiency between two detection configurations emitting the gamma ray in different energies assuming a "virtual" reference point detection configuration. No pre-optimization of the detector geometrical characteristics was performed before the transfer to test the ability of the efficiency transfer to reduce the effect of the ignorance on their real magnitude on the quality of the transferred efficiency. The obtained and measured efficiencies were found in good agreement for the two investigated methods of efficiency transfer. The obtained agreement proves that Monte Carlo method and especially the GEANT4 code constitute an efficient tool to obtain accurate detection efficiency values. The second investigated efficiency transfer procedure is useful to calibrate the HPGe gamma detector for any emission energy value for a voluminous source using one point source detection efficiency emitting in a different energy as a reference efficiency. The calculations preformed in this work were applied to the measurement exercise of the EUROMET428 project. A measurement exercise where an evaluation of the full energy peak efficiencies in the energy range 60-2000 keV for a typical coaxial p-type HpGe detector and several types of source configuration: point sources located at various distances from the detector and a cylindrical box containing three matrices was performed.

  17. Tectonic calibrations in molecular dating

    Institute of Scientific and Technical Information of China (English)

    Ullasa KODANDARAMAIAH

    2011-01-01

    Molecular dating techniques require the use of calibrations, which are usually fossil or geological vicariance-based.Fossil calibrations have been criticised because they result only in minimum age estimates. Based on a historical biogeographic perspective, Ⅰ suggest that vicariance-based calibrations are more dangerous. Almost all analytical methods in historical biogeography are strongly biased towards inferring vicariance, hence vicariance identified through such methods is unreliable. Other studies, especially of groups found on Gondwanan fragments, have simply assumed vicariance. Although it was previously believed that vicariance was the predominant mode of speciation, mounting evidence now indicates that speciation by dispersal is common, dominating vicariance in several groups. Moreover, the possibility of speciation having occurred before the said geological event cannot be precluded. Thus, geological calibrations can under- or overestimate times, whereas fossil calibrations always result in minimum estimates. Another major drawback of vicariant calibrations is the problem of circular reasoning when the resulting estimates are used to infer ages of biogeographic events. Ⅰ argue that fossil-based dating is a superior alternative to vicariance, primarily because the strongest assumption in the latter, that speciation was caused by the said geological process, is more often than not the most tenuous. When authors prefer to use a combination of fossil and vicariant calibrations, one suggestion is to report results both with and without inclusion of the geological constraints. Relying solely on vicariant calibrations should be strictly avoided.

  18. Cobalt source calibration

    Energy Technology Data Exchange (ETDEWEB)

    Rizvi, H.M.

    1999-12-03

    The data obtained from these tests determine the dose rate of the two cobalt sources in SRTC. Building 774-A houses one of these sources while the other resides in room C-067 of Building 773-A. The data from this experiment shows the following: (1) The dose rate of the No.2 cobalt source in Building 774-A measured 1.073 x 10{sup 5} rad/h (June 17, 1999). The dose rate of the Shepherd Model 109 Gamma cobalt source in Building 773-A measured 9.27 x 10{sup 5} rad/h (June 25, 1999). These rates come from placing the graduated cylinder containing the dosimeter solution in the center of the irradiation chamber. (2) Two calibration tests in the 774-A source placed the graduated cylinder with the dosimeter solution approximately 1.5 inches off center in the axial direction. This movement of the sample reduced the measured dose rate 0.92% from 1.083 x 10{sup 5} rad/h to 1.073 x 10{sup 5} rad/h. and (3) A similar test in the cobalt source in 773-A placed the graduated cylinder approximately 2.0 inches off center in the axial direction. This change in position reduced the measured dose rate by 10.34% from 1.036 x 10{sup 6} to 9.27 x 10{sup 5}. This testing used chemical dosimetry to measure the dose rate of a radioactive source. In this method, one determines the dose by the chemical change that takes place in the dosimeter. For this calibration experiment, the author used a Fricke (ferrous ammonium sulfate) dosimeter. This solution works well for dose rates to 10{sup 7} rad/h. During irradiation of the Fricke dosimeter solution the Fe{sup 2+} ions ionize to Fe{sup 3+}. When this occurs, the solution acquires a slightly darker tint (not visible to the human eye). To determine the magnitude of the change in Fe ions, one places the solution in an UV-VIS Spectrophotometer. The UV-VIS Spectrophotometer measures the absorbency of the solution. Dividing the absorbency by the total time (in minutes) of exposure yields the dose rate.

  19. The Advanced LIGO Photon Calibrators

    CERN Document Server

    Karki, S; Kandhasamy, S; Abbott, B P; Abbott, T D; Anders, E H; Berliner, J; Betzwieser, J; Daveloza, H P; Cahillane, C; Canete, L; Conley, C; Gleason, J R; Goetz, E; Kissel, J S; Izumi, K; Mendell, G; Quetschke, V; Rodruck, M; Sachdev, S; Sadecki, T; Schwinberg, P B; Sottile, A; Wade, M; Weinstein, A J; West, M; Savage, R L

    2016-01-01

    The two interferometers of the Laser Interferometry Gravitaional-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events, and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as Photon Calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO Photon Calibrators that are currently providing fiducial displacements on the order of $10^{-18}$ m/$\\sqrt{\\textrm{Hz}}$ with accuracy and precision of better ...

  20. The Advanced LIGO photon calibrators

    Science.gov (United States)

    Karki, S.; Tuyenbayev, D.; Kandhasamy, S.; Abbott, B. P.; Abbott, T. D.; Anders, E. H.; Berliner, J.; Betzwieser, J.; Cahillane, C.; Canete, L.; Conley, C.; Daveloza, H. P.; De Lillo, N.; Gleason, J. R.; Goetz, E.; Izumi, K.; Kissel, J. S.; Mendell, G.; Quetschke, V.; Rodruck, M.; Sachdev, S.; Sadecki, T.; Schwinberg, P. B.; Sottile, A.; Wade, M.; Weinstein, A. J.; West, M.; Savage, R. L.

    2016-11-01

    The two interferometers of the Laser Interferometry Gravitational-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as photon calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO photon calibrators that are currently providing fiducial displacements on the order of 1 0-18m /√{Hz } with accuracy and precision of better than 1%.

  1. TIME CALIBRATED OSCILLOSCOPE SWEEP CIRCUIT

    Science.gov (United States)

    Smith, V.L.; Carstensen, H.K.

    1959-11-24

    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  2. Automated calibration of multistatic arrays

    Energy Technology Data Exchange (ETDEWEB)

    Henderer, Bruce

    2017-03-14

    A method is disclosed for calibrating a multistatic array having a plurality of transmitter and receiver pairs spaced from one another along a predetermined path and relative to a plurality of bin locations, and further being spaced at a fixed distance from a stationary calibration implement. A clock reference pulse may be generated, and each of the transmitters and receivers of each said transmitter/receiver pair turned on at a monotonically increasing time delay interval relative to the clock reference pulse. Ones of the transmitters and receivers may be used such that a previously calibrated transmitter or receiver of a given one of the transmitter/receiver pairs is paired with a subsequently un-calibrated one of the transmitters or receivers of an immediately subsequently positioned transmitter/receiver pair, to calibrate the transmitter or receiver of the immediately subsequent transmitter/receiver pair.

  3. Antenna Calibration and Measurement Equipment

    Science.gov (United States)

    Rochblatt, David J.; Cortes, Manuel Vazquez

    2012-01-01

    A document describes the Antenna Calibration & Measurement Equipment (ACME) system that will provide the Deep Space Network (DSN) with instrumentation enabling a trained RF engineer at each complex to perform antenna calibration measurements and to generate antenna calibration data. This data includes continuous-scan auto-bore-based data acquisition with all-sky data gathering in support of 4th order pointing model generation requirements. Other data includes antenna subreflector focus, system noise temperature and tipping curves, antenna efficiency, reports system linearity, and instrument calibration. The ACME system design is based on the on-the-fly (OTF) mapping technique and architecture. ACME has contributed to the improved RF performance of the DSN by approximately a factor of two. It improved the pointing performances of the DSN antennas and productivity of its personnel and calibration engineers.

  4. A calibrated Franklin chimes

    Science.gov (United States)

    Gonta, Igor; Williams, Earle

    1994-05-01

    Benjamin Franklin devised a simple yet intriguing device to measure electrification in the atmosphere during conditions of foul weather. He constructed a system of bells, one of which was attached to a conductor that was suspended vertically above his house. The device is illustrated in a well-known painting of Franklin (Cohen, 1985). The elevated conductor acquired a potential due to the electric field in the atmosphere and caused a brass ball to oscillate between two bells. The purpose of this study is to extend Franklin's idea by constructing a set of 'chimes' which will operate both in fair and in foul weather conditions. In addition, a mathematical relationship will be established between the frequency of oscillation of a metallic sphere in a simplified geometry and the potential on one plate due to the electrification of the atmosphere. Thus it will be possible to calibrate the 'Franklin Chimes' and to obtain a nearly instantaneous measurement of the potential of the elevated conductor in both fair and foul weather conditions.

  5. Standard practice for torque calibration of testing machines and devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This practice covers procedures and requirements for the calibration of torque for static and quasi-static torque capable testing machines or devices. These may, or may not, have torque indicating systems and include those devices used for the calibration of hand torque tools. Testing machines may be calibrated by one of the three following methods or combination thereof: 1.1.1 Use of standard weights and lever arms. 1.1.2 Use of elastic torque measuring devices. 1.1.3 Use of elastic force measuring devices and lever arms. 1.1.4 Any of the methods require a specific uncertainty of measurement and a traceability derived from national standards of mass and length. 1.2 The procedures of 1.1.1, 1.1.2, and 1.1.3 apply to the calibration of the torque-indicating systems associated with the testing machine, such as a scale, dial, marked or unmarked recorder chart, digital display, etc. In all cases the buyer/owner/user must designate the torque-indicating system(s) to be calibrated and included in the repor...

  6. [VMTBB-Based Spectral Radiometric Calibration of NIR Fiber Coupled Spectrometer].

    Science.gov (United States)

    Zheng, Feng; Liu, Li-ying; Liu, Xiao-xi; Li, Ye; Shi, Xiao-guang; Zhang, Guo-yu; Huan, Ke-wei

    2015-09-01

    The medium temperature black body (MTBB) is conventional high precision equipment used as spectral radiometric scale in infrared spectral region. However, in near-infrared (NIR) spectral region, there are few papers about spectral radiometric calibration by using MTBB, that is because NIR spectral region is the borderland of its effective spectral region. The main research of this paper is spectral radiometric calibration method by using MTBB in NIR spectral region. Accordingly, this paper is devoted mostly to a discussion of how the calibration precision could be affected by selecting different structural parameters of calibration model. The purpose of this paper is to present the results of research and provide technical reference for improving the traceability in NIR spectral radiometric calibration. In this paper, a NIR fiber coupled spectrometer, whose wavelength range covers from 950 to 1700 nm, has been calibrated by a MTBB with adjustable temperature range from 50 to 1050 °C. Concentrating on calibration process, two key points have been discussed. For one thing, the geometric factors of radiation transfer model of the calibration systems have been compared between traditional structure and fiber direct-coupled structure. Because the fiber direct-coupled model is simple and effective, it has been selected instead of traditional model based on the radiation transfer between two coaxial discs. So, it is an advantaged radiation transfer model for radiometric calibration of fiber coupled spectrometer. For another thing, the relation between calibration accuracy and structural parameters of calibration model has been analyzed intensively. The root cause is scale feature of attribute of calibration data itself, which is the nonlinear structure in scales of spectral data. So, the high precision calibration needs nonlinear calibration model, and the uniform sampling for scale feature is also very important. Selecting sample is an inevitable problem when the

  7. LBCS: The LOFAR Long-Baseline Calibrator Survey

    Science.gov (United States)

    Jackson, N.; Tagore, A.; Deller, A.; Moldón, J.; Varenius, E.; Morabito, L.; Wucknitz, O.; Carozzi, T.; Conway, J.; Drabent, A.; Kapinska, A.; Orrù, E.; Brentjens, M.; Blaauw, R.; Kuper, G.; Sluman, J.; Schaap, J.; Vermaas, N.; Iacobelli, M.; Cerrigone, L.; Shulevski, A.; ter Veen, S.; Fallows, R.; Pizzo, R.; Sipior, M.; Anderson, J.; Avruch, I. M.; Bell, M. E.; van Bemmel, I.; Bentum, M. J.; Best, P.; Bonafede, A.; Breitling, F.; Broderick, J. W.; Brouw, W. N.; Brüggen, M.; Ciardi, B.; Corstanje, A.; de Gasperin, F.; de Geus, E.; Eislöffel, J.; Engels, D.; Falcke, H.; Garrett, M. A.; Grießmeier, J. M.; Gunst, A. W.; van Haarlem, M. P.; Heald, G.; Hoeft, M.; Hörandel, J.; Horneffer, A.; Intema, H.; Juette, E.; Kuniyoshi, M.; van Leeuwen, J.; Loose, G. M.; Maat, P.; McFadden, R.; McKay-Bukowski, D.; McKean, J. P.; Mulcahy, D. D.; Munk, H.; Pandey-Pommier, M.; Polatidis, A. G.; Reich, W.; Röttgering, H. J. A.; Rowlinson, A.; Scaife, A. M. M.; Schwarz, D. J.; Steinmetz, M.; Swinbank, J.; Thoudam, S.; Toribio, M. C.; Vermeulen, R.; Vocks, C.; van Weeren, R. J.; Wise, M. W.; Yatawatta, S.; Zarka, P.

    2016-11-01

    We outline the LOFAR Long-Baseline Calibrator Survey (LBCS), whose aim is to identify sources suitable for calibrating the highest-resolution observations made with the International LOFAR Telescope, which include baselines >1000 km. Suitable sources must contain significant correlated flux density (≳ 50 - 100 mJy) at frequencies around 110-190 MHz on scales of a few hundred milliarcseconds. At least for the 200-300-km international baselines, we find around 1 suitable calibrator source per square degree over a large part of the northern sky, in agreement with previous work. This should allow a randomly selected target to be successfully phase calibrated on the international baselines in over 50% of cases. Products of the survey include calibrator source lists and fringe-rate and delay maps of wide areas - typically a few degrees - around each source. The density of sources with significant correlated flux declines noticeably with baseline length over the range 200-600 km, with good calibrators on the longest baselines appearing only at the rate of 0.5 per sq. deg. Coherence times decrease from 1-3 min on 200-km baselines to about 1 min on 600-km baselines, suggesting that ionospheric phase variations contain components with scales of a few hundred kilometres. The longest median coherence time, at just over 3 min, is seen on the DE609 baseline, which at 227 km is close to being the shortest. We see median coherence times of between 80 and 110 s on the four longest baselines (580-600 km), and about 2 min for the other baselines. The success of phase transfer from calibrator to target is shown to be influenced by distance, in a manner that suggests a coherence patch at 150-MHz of the order of 1 deg. Although source structures cannot be measured in these observations, we deduce that phase transfer is affected if the calibrator source structure is not known. We give suggestions for calibration strategies and choice of calibrator sources, and describe the access to

  8. Radionuclide transfer. Radionuklid Transfer

    Energy Technology Data Exchange (ETDEWEB)

    Gerber, G.B.

    1993-01-01

    The research project described here had the aim to obtain further information on the transfer of nuclides during pregnancy and lactation. The tests were carried out in mini-pigs and rats receiving unchanging doses of radionuclides with the food. The following findings were revealed for the elements examined: Fe, Se, Cs and Zn were characterized by very high transfer levels in the mother, infant and foetus. A substantial uptake by the mother alone was observed for Co, Ag and Mn. The uptake by the foetus and infant here was 1 to 10 times lower. A preferential concentration in certain tissues was seen for Sr and Tc; the thyroid levels of Tc were about equally high in mothers and infants, while Sr showed less accumulation in the maternal bone. The lanthanide group of substances (Ce, Eu and Gd as well as Y and Ru) were only taken up to a very limited extent. The uptake of the examined radionuclides (Fe, Co, Ag, Ce) with the food ingested was found here to be ten times greater in rats as compared to mini-pigs. This showed that great caution must be observed, if the behaviour of radionuclides in man is extrapolated from relevant data obtained in rodents. (orig./MG)

  9. 成熟卵泡获卵率对体外受精结局的影响%The influence of retrieved oocyte rate of mature follicles on the outcomes during in vitro fertilization & embryo transfer procedure

    Institute of Scientific and Technical Information of China (English)

    马黔红; 黄仲英; 樊伟

    2011-01-01

    Objective To investigate the influence of retrieved oocyte rate on clinical and laboratory results during in vitro fertilization and embryo transfer or intracytoplasmic sperm injection procedure. Methods 746 IVF/ ICSI cycles (693 patients) were analysed retrospectively in West China Second University Hospital of Sichuan University from May to December in 2010. Based on retrieved oocyte rate, all these cycles were divided into four groups: group A, retrieved oocyte rate ≥80%; group B, retrieved oocyte rate 50 %~79%; group C, retrieved o-ocyte rate 30%~49%; group D, retrieved oocyte rate 0. 05). The cases from group D obtained the lowest number of retrieved oocytes in the four groups, and there were significantly lower in group D than in group A and B in maturation rate of oocytes, fertilization rate, cleavage rate and embryo formation rate that could be transferred (P< 0. 05) , which resulting to the lower clinical pregnancy rate, implantation rate and cumulative pregnancy rate in group D than in group A and B (P<0. 05). Also, there were significantly lower in fertilization rate, cleavage rate, embryo formation rate that could be transferred in group C than in group A (P<0. 05), but cumulative pregnancy rate in group C was similar to group A and B. Conclusions Decreased retrieved oocyte rate may cause decreased oocyte maturation rate, fertilization rate, cleavage rate and good quality embryo rate, so as to lessen the clinical pregnancy rate.%目的 比较在体外受精-胚胎移植/卵细胞浆内单精子注射(IVF- ET/ICSI)过程中,控制性超促排卵(COH)后成熟卵泡获卵率对IVF- ET/ICSI的实验室和临床结局的影响.方法 回顾性分析2010年5月至2010年12月四川大学华西第二医院行IVF- ET/ICSI治疗的693例患者,共746个周期,根据获卵率将患者分为A组:获卵率≥80%;B组:获卵率50%~79%;C组:获卵率30%~49%;D组:获卵率<30%.比较4组患者的临床和实验室结局.结果 A、B组卵子成

  10. Analysis of Nigeria research reactor-1 thermal power calibration methods

    Energy Technology Data Exchange (ETDEWEB)

    Agbo, Sunday Arome; Ahmed, Yusuf Aminu; Ewa, Ita Okon; Jibrin, Yahaya [Ahmadu Bello University, Zaria (Nigeria)

    2016-06-15

    This paper analyzes the accuracy of the methods used in calibrating the thermal power of Nigeria Research Reactor-1 (NIRR-1), a low-power miniature neutron source reactor located at the Centre for Energy Research and Training, Ahmadu Bello University, Zaria, Nigeria. The calibration was performed at three different power levels: low power (3.6 kW), half power (15 kW), and full power (30 kW). Two methods were used in the calibration, namely, slope and heat balance methods. The thermal power obtained by the heat balance method at low power, half power, and full power was 3.7 ± 0.2 kW, 15.2 ± 1.2 kW, and 30.7 ± 2.5 kW, respectively. The thermal power obtained by the slope method at half power and full power was 15.8 ± 0.7 kW and 30.2 ± 1.5 kW, respectively. It was observed that the slope method is more accurate with deviations of 4% and 5% for calibrations at half and full power, respectively, although the linear fit (slope method) on average temperature-rising rates during the thermal power calibration procedure at low power (3.6 kW) is not fitting. As such, the slope method of power calibration is not suitable at lower power for NIRR-1.

  11. Data filtering with support vector machines in geometric camera calibration.

    Science.gov (United States)

    Ergun, B; Kavzoglu, T; Colkesen, I; Sahin, C

    2010-02-01

    The use of non-metric digital cameras in close-range photogrammetric applications and machine vision has become a popular research agenda. Being an essential component of photogrammetric evaluation, camera calibration is a crucial stage for non-metric cameras. Therefore, accurate camera calibration and orientation procedures have become prerequisites for the extraction of precise and reliable 3D metric information from images. The lack of accurate inner orientation parameters can lead to unreliable results in the photogrammetric process. A camera can be well defined with its principal distance, principal point offset and lens distortion parameters. Different camera models have been formulated and used in close-range photogrammetry, but generally sensor orientation and calibration is performed with a perspective geometrical model by means of the bundle adjustment. In this study, support vector machines (SVMs) using radial basis function kernel is employed to model the distortions measured for Olympus Aspherical Zoom lens Olympus E10 camera system that are later used in the geometric calibration process. It is intended to introduce an alternative approach for the on-the-job photogrammetric calibration stage. Experimental results for DSLR camera with three focal length settings (9, 18 and 36 mm) were estimated using bundle adjustment with additional parameters, and analyses were conducted based on object point discrepancies and standard errors. Results show the robustness of the SVMs approach on the correction of image coordinates by modelling total distortions on-the-job calibration process using limited number of images.

  12. Inter-calibration of polar imager solar channels using SEVIRI

    Directory of Open Access Journals (Sweden)

    J. F. Meirink

    2013-04-01

    Full Text Available Accurate calibration of satellite imagers is a prerequisite for using their measurements in climate applications. Here we present a method for the inter-calibration of geostationary and polar-orbiting imager solar channels based on regressions of collocated near-nadir radiances. Specific attention is paid to correcting for differences in spectral response between instruments. The method is used to calibrate the solar channels of the Spinning Enhanced Visible and Infrared Imager (SEVIRI on the geostationary Meteosat satellite with corresponding channels of the Moderate Resolution Imaging Spectroradiometer (MODIS on the polar-orbiting Aqua satellite. The SEVIRI operational calibration is found to be stable during the years 2004 to 2009 but off by −8, −6, and +3.5% for channels 1 (0.6 μm, 2 (0.8 μm, and 3 (1.6 μm, respectively. These results are robust for a range of choices that can be made regarding data collocation and selection, as long as the viewing and illumination geometries of the two instruments are matched. Uncertainties in the inter-calibration method are estimated to be 1% for channel 1 and 1.5% for channels 2 and 3. A specific application of the method is the inter-calibration of polar imagers using SEVIRI as a transfer instrument. This offers an alternative to direct inter-calibration, which in general has to rely on high-latitude collocations. Using this method we have tied MODIS-Terra and Advanced Very High Resolution Radiometer (AVHRR instruments on National Oceanic and Atmospheric Administration (NOAA satellites 17 and 18 to MODIS-Aqua for the years 2007 to 2009. While reflectances of the two MODIS instruments differ less than 2% for all channels considered, deviations of an existing AVHRR calibration from MODIS-Aqua reach −3.5 and +2.5% for the 0.8 and 1.6 μm channels, respectively.

  13. Novel real-time alignment and calibration of the LHCb detector in Run II

    CERN Document Server

    Xu, Zhirui

    2015-01-01

    An automatic real-time alignment and calibration strategy of the LHCb detector was developed for the Run II. Thanks to the online calibration, tighter event selection criteria can be used in the trigger. Furthermore, the online calibration facilitates the use of hadronic particle identification using the Ring Imaging Cherenkov (RICH) detectors at the trigger level. The motivation for a real-time alignment and calibration of the LHCb detector is discussed from both the operational and physics performance points of view. Specific challenges of this novel configuration are discussed, as well as the working procedures of the framework and its performance.

  14. Calibration of the linear optics in the SSRF storage ring

    Institute of Scientific and Technical Information of China (English)

    TIAN Shunqiang; LIU Guimin; ZHANG Wenzhi; LI Haohu; ZHANG Manzhou; HOU Jie; CHEN Guangling

    2008-01-01

    Phase Ⅰ commissioning of the SSRF storage ring at 3.0 GeV was ended with encouraging results. Distortions and calibrations of the linear optics during the storage ring commissioning are discussed in this paper. The calibration procedure has reduced sextupole-additional focusing effects by minimizing closed orbit deviation and corrected quadrupole magnetic coefficients with the linear optics from closed orbit (LOCO) technique. After fitting the closed orbit response matrix, linear optics of the storage ring is substantially corrected, and the measured parameters agree well with the design. Four optics modes were commissioned, and relevant machine physics studies were carried out. Their results are summarized.

  15. Auto-calibration of Systematic Odometry Errors in Mobile Robots

    DEFF Research Database (Denmark)

    Bak, Martin; Larsen, Thomas Dall; Andersen, Nils Axel;

    1999-01-01

    This paper describes the phenomenon of systematic errors in odometry models in mobile robots and looks at various ways of avoiding it by means of auto-calibration. The systematic errors considered are incorrect knowledge of the wheel base and the gains from encoder readings to wheel displacement....... By auto-calibration we mean a standardized procedure which estimates the uncertainties using only on-board equipment such as encoders, an absolute measurement system and filters; no intervention by operator or off-line data processing is necessary. Results are illustrated by a number of simulations...

  16. A new method for thread calibration on coordinate measuring machines

    DEFF Research Database (Denmark)

    Carmignato, Simone; De Chiffre, Leonardo

    2003-01-01

    CIRP Annals – Paper proposal temporary reference: P15. This paper presents a new method for the calibration of thread gauges on coordinate measuring machines. The procedure involves scanning of thread profiles using a needle-like probe, achieving traceability by substitution of different thread......-3 gave measuring uncertainties comparable to the values from usual calibration methods on dedicated equipment, e.g. a measuring uncertainty of 1.5 µm was achieved for measurement of the pitch, and 2-2.5 µm for diameter measurements....

  17. Challenges in X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren; Rasmussen, Michael R.

    2009-01-01

    Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis and real time control purposes. In these contexts, it is all-important that the radar data well calibrated and adjusted in order to obtain valid quantitative precipitation...... estimates. This paper compares two calibration procedures for a small marine X-band radar by comparing radar data with rain gauge data. Validation shows a very good consensus with regards to precipitation volumes, but more diverse results on peak rain intensities....

  18. Calibrating an interferometric laser frequency stabilization to MHz precision

    CERN Document Server

    Brachmann, Johannes F S; Dieckmann, Kai; 10.1364/AO.51.005517

    2012-01-01

    We report on a calibration procedure that enhances the precision of an interferometer based frequency stabilization by several orders of magnitude. For this purpose the frequency deviations of the stabilization are measured precisely by means of a frequency comb. This allows to implement several calibration steps that compensate different systematic errors. The resulting frequency deviation is shown to be less than $5.7 $MHz (rms $1.6 $MHz) in the whole wavelength interval $750 - 795 $nm. Wide tuning of a stabilized laser at this exceptional precision is demonstrated.

  19. Jet energy calibration in ATLAS

    CERN Document Server

    Schouten, Doug

    A correct energy calibration for jets is essential to the success of the ATLAS experi- ment. In this thesis I study a method for deriving an in situ jet energy calibration for the ATLAS detector. In particular, I show the applicability of the missing transverse energy projection fraction method. This method is shown to set the correct mean energy for jets. Pileup effects due to the high luminosities at ATLAS are also stud- ied. I study the correlations in lateral distributions of pileup energy, as well as the luminosity dependence of the in situ calibration metho

  20. Calibrating System for Vacuum Gauges

    Institute of Scientific and Technical Information of China (English)

    MengJun; YangXiaotian; HaoBinggan; HouShengjun; HuZhenjun

    2003-01-01

    In order to measure the vacuum degree, a lot of vacuum gauges will be used in CSR vacuum system. We bought several types of vacuum gauges. We know that different typos of vacuum gauges or even one type of vacuum gauges have different measure results in same condition, so they must be calibrated. But it seems impossible for us to send so many gauges to the calibrating station outside because of the high price. So the best choice is to build a second class calibrating station for vacuum gauges by ourselves (Fig.l).

  1. Calibration of the CMS Drift Tube Chambers and Measurement of the Drift Velocity with Cosmic Rays

    CERN Document Server

    Chatrchyan, S; Sirunyan, A M; Adam, W; Arnold, B; Bergauer, H; Bergauer, T; Dragicevic, M; Eichberger, M; Erö, J; Friedl, M; Frühwirth, R; Ghete, V M; Hammer, J; Hänsel, S; Hoch, M; Hörmann, N; Hrubec, J; Jeitler, M; Kasieczka, G; Kastner, K; Krammer, M; Liko, D; Magrans de Abril, I; Mikulec, I; Mittermayr, F; Neuherz, B; Oberegger, M; Padrta, M; Pernicka, M; Rohringer, H; Schmid, S; Schöfbeck, R; Schreiner, T; Stark, R; Steininger, H; Strauss, J; Taurok, A; Teischinger, F; Themel, T; Uhl, D; Wagner, P; Waltenberger, W; Walzel, G; Widl, E; Wulz, C E; Chekhovsky, V; Dvornikov, O; Emeliantchik, I; Litomin, A; Makarenko, V; Marfin, I; Mossolov, V; Shumeiko, N; Solin, A; Stefanovitch, R; Suarez Gonzalez, J; Tikhonov, A; Fedorov, A; Karneyeu, A; Korzhik, M; Panov, V; Zuyeuski, R; Kuchinsky, P; Beaumont, W; Benucci, L; Cardaci, M; De Wolf, E A; Delmeire, E; Druzhkin, D; Hashemi, M; Janssen, X; Maes, T; Mucibello, L; Ochesanu, S; Rougny, R; Selvaggi, M; Van Haevermaet, H; Van Mechelen, P; Van Remortel, N; Adler, V; Beauceron, S; Blyweert, S; D'Hondt, J; De Weirdt, S; Devroede, O; Heyninck, J; Kalogeropoulos, A; Maes, J; Maes, M; Mozer, M U; Tavernier, S; Van Doninck, W; Van Mulders, P; Villella, I; Bouhali, O; Chabert, E C; Charaf, O; Clerbaux, B; De Lentdecker, G; Dero, V; Elgammal, S; Gay, A P R; Hammad, G H; Marage, P E; Rugovac, S; Vander Velde, C; Vanlaer, P; Wickens, J; Grunewald, M; Klein, B; Marinov, A; Ryckbosch, D; Thyssen, F; Tytgat, M; Vanelderen, L; Verwilligen, P; Basegmez, S; Bruno, G; Caudron, J; Delaere, C; Demin, P; Favart, D; Giammanco, A; Grégoire, G; Lemaitre, V; Militaru, O; Ovyn, S; Piotrzkowski, K; Quertenmont, L; Schul, N; Beliy, N; Daubie, E; Alves, G A; Pol, M E; Souza, M H G; Carvalho, W; De Jesus Damiao, D; De Oliveira Martins, C; Fonseca De Souza, S; Mundim, L; Oguri, V; Santoro, A; Silva Do Amaral, S M; Sznajder, A; Fernandez Perez Tomei, T R; Ferreira Dias, M A; Gregores, E M; Novaes, S F; Abadjiev, K; Anguelov, T; Damgov, J; Darmenov, N; Dimitrov, L; Genchev, V; Iaydjiev, P; Piperov, S; Stoykova, S; Sultanov, G; Trayanov, R; Vankov, I; Dimitrov, A; Dyulendarova, M; Kozhuharov, V; Litov, L; Marinova, E; Mateev, M; Pavlov, B; Petkov, P; Toteva, Z; Chen, G M; Chen, H S; Guan, W; Jiang, C H; Liang, D; Liu, B; Meng, X; Tao, J; Wang, J; Wang, Z; Xue, Z; Zhang, Z; Ban, Y; Cai, J; Ge, Y; Guo, S; Hu, Z; Mao, Y; Qian, S J; Teng, H; Zhu, B; Avila, C; Baquero Ruiz, M; Carrillo Montoya, C A; Gomez, A; Gomez Moreno, B; Ocampo Rios, A A; Osorio Oliveros, A F; Reyes Romero, D; Sanabria, J C; Godinovic, N; Lelas, K; Plestina, R; Polic, D; Puljak, I; Antunovic, Z; Dzelalija, M; Brigljevic, V; Duric, S; Kadija, K; Morovic, S; Fereos, R; Galanti, M; Mousa, J; Papadakis, A; Ptochos, F; Razis, P A; Tsiakkouri, D; Zinonos, Z; Hektor, A; Kadastik, M; Kannike, K; Müntel, M; Raidal, M; Rebane, L; Anttila, E; Czellar, S; Härkönen, J; Heikkinen, A; Karimäki, V; Kinnunen, R; Klem, J; Kortelainen, M J; Lampén, T; Lassila-Perini, K; Lehti, S; Lindén, T; Luukka, P; Mäenpää, T; Nysten, J; Tuominen, E; Tuominiemi, J; Ungaro, D; Wendland, L; Banzuzi, K; Korpela, A; Tuuva, T; Nedelec, P; Sillou, D; Besancon, M; Chipaux, R; Dejardin, M; Denegri, D; Descamps, J; Fabbro, B; Faure, J L; Ferri, F; Ganjour, S; Gentit, F X; Givernaud, A; Gras, P; Hamel de Monchenault, G; Jarry, P; Lemaire, M C; Locci, E; Malcles, J; Marionneau, M; Millischer, L; Rander, J; Rosowsky, A; Rousseau, D; Titov, M; Verrecchia, P; Baffioni, S; Bianchini, L; Bluj, M; Busson, P; Charlot, C; Dobrzynski, L; Granier de Cassagnac, R; Haguenauer, M; Miné, P; Paganini, P; Sirois, Y; Thiebaux, C; Zabi, A; Agram, J L; Besson, A; Bloch, D; Bodin, D; Brom, J M; Conte, E; Drouhin, F; Fontaine, J C; Gelé, D; Goerlach, U; Gross, L; Juillot, P; Le Bihan, A C; Patois, Y; Speck, J; Van Hove, P; Baty, C; Bedjidian, M; Blaha, J; Boudoul, G; Brun, H; Chanon, N; Chierici, R; Contardo, D; Depasse, P; Dupasquier, T; El Mamouni, H; Fassi, F; Fay, J; Gascon, S; Ille, B; Kurca, T; Le Grand, T; Lethuillier, M; Lumb, N; Mirabito, L; Perries, S; Vander Donckt, M; Verdier, P; Djaoshvili, N; Roinishvili, N; Roinishvili, V; Amaglobeli, N; Adolphi, R; Anagnostou, G; Brauer, R; Braunschweig, W; Edelhoff, M; Esser, H; Feld, L; Karpinski, W; Khomich, A; Klein, K; Mohr, N; Ostaptchouk, A; Pandoulas, D; Pierschel, G; Raupach, F; Schael, S; Schultz von Dratzig, A; Schwering, G; Sprenger, D; Thomas, M; Weber, M; Wittmer, B; Wlochal, M; Actis, O; Altenhöfer, G; Bender, W; Biallass, P; Erdmann, M; Fetchenhauer, G; Frangenheim, J; Hebbeker, T; Hilgers, G; Hinzmann, A; Hoepfner, K; Hof, C; Kirsch, M; Klimkovich, T; Kreuzer, P; Lanske, D; Merschmeyer, M; Meyer, A; Philipps, B; Pieta, H; Reithler, H; Schmitz, S A; Sonnenschein, L; Sowa, M; Steggemann, J; Szczesny, H; Teyssier, D; Zeidler, C; Bontenackels, M; Davids, M; Duda, M; Flügge, G; Geenen, H; Giffels, M; Haj Ahmad, W; Hermanns, T; Heydhausen, D; Kalinin, S; Kress, T; Linn, A; Nowack, A; Perchalla, L; Poettgens, M; Pooth, O; Sauerland, P; Stahl, A; Tornier, D; Zoeller, M H; Aldaya Martin, M; Behrens, U; Borras, K; Campbell, A; Castro, E; Dammann, D; Eckerlin, G; Flossdorf, A; Flucke, G; Geiser, A; Hatton, D; Hauk, J; Jung, H; Kasemann, M; Katkov, I; Kleinwort, C; Kluge, H; Knutsson, A; Kuznetsova, E; Lange, W; Lohmann, W; Mankel, R; Marienfeld, M; Meyer, A B; Miglioranzi, S; Mnich, J; Ohlerich, M; Olzem, J; Parenti, A; Rosemann, C; Schmidt, R; Schoerner-Sadenius, T; Volyanskyy, D; Wissing, C; Zeuner, W D; Autermann, C; Bechtel, F; Draeger, J; Eckstein, D; Gebbert, U; Kaschube, K; Kaussen, G; Klanner, R; Mura, B; Naumann-Emme, S; Nowak, F; Pein, U; Sander, C; Schleper, P; Schum, T; Stadie, H; Steinbrück, G; Thomsen, J; Wolf, R; Bauer, J; Blüm, P; Buege, V; Cakir, A; Chwalek, T; De Boer, W; Dierlamm, A; Dirkes, G; Feindt, M; Felzmann, U; Frey, M; Furgeri, A; Gruschke, J; Hackstein, C; Hartmann, F; Heier, S; Heinrich, M; Held, H; Hirschbuehl, D; Hoffmann, K H; Honc, S; Jung, C; Kuhr, T; Liamsuwan, T; Martschei, D; Mueller, S; Müller, Th; Neuland, M B; Niegel, M; Oberst, O; Oehler, A; Ott, J; Peiffer, T; Piparo, D; Quast, G; Rabbertz, K; Ratnikov, F; Ratnikova, N; Renz, M; Saout, C; Sartisohn, G; Scheurer, A; Schieferdecker, P; Schilling, F P; Schott, G; Simonis, H J; Stober, F M; Sturm, P; Troendle, D; Trunov, A; Wagner, W; Wagner-Kuhr, J; Zeise, M; Zhukov, V; Ziebarth, E B; Daskalakis, G; Geralis, T; Karafasoulis, K; Kyriakis, A; Loukas, D; Markou, A; Markou, C; Mavrommatis, C; Petrakou, E; Zachariadou, A; Gouskos, L; Katsas, P; Panagiotou, A; Evangelou, I; Kokkas, P; Manthos, N; Papadopoulos, I; Patras, V; Triantis, F A; Bencze, G; Boldizsar, L; Debreczeni, G; Hajdu, C; Hernath, S; Hidas, P; Horvath, D; Krajczar, K; Laszlo, A; Patay, G; Sikler, F; Toth, N; Vesztergombi, G; Beni, N; Christian, G; Imrek, J; Molnar, J; Novak, D; Palinkas, J; Szekely, G; Szillasi, Z; Tokesi, K; Veszpremi, V; Kapusi, A; Marian, G; Raics, P; Szabo, Z; Trocsanyi, Z L; Ujvari, B; Zilizi, G; Bansal, S; Bawa, H S; Beri, S B; Bhatnagar, V; Jindal, M; Kaur, M; Kaur, R; Kohli, J M; Mehta, M Z; Nishu, N; Saini, L K; Sharma, A; Singh, A; Singh, J B; Singh, S P; Ahuja, S; Arora, S; Bhattacharya, S; Chauhan, S; Choudhary, B C; Gupta, P; Jain, S; Jain, S; Jha, M; Kumar, A; Ranjan, K; Shivpuri, R K; Srivastava, A K; Choudhury, R K; Dutta, D; Kailas, S; Kataria, S K; Mohanty, A K; Pant, L M; Shukla, P; Topkar, A; Aziz, T; Guchait, M; Gurtu, A; Maity, M; Majumder, D; Majumder, G; Mazumdar, K; Nayak, A; Saha, A; Sudhakar, K; Banerjee, S; Dugad, S; Mondal, N K; Arfaei, H; Bakhshiansohi, H; Fahim, A; Jafari, A; Mohammadi Najafabadi, M; Moshaii, A; Paktinat Mehdiabadi, S; Rouhani, S; Safarzadeh, B; Zeinali, M; Felcini, M; Abbrescia, M; Barbone, L; Chiumarulo, F; Clemente, A; Colaleo, A; Creanza, D; Cuscela, G; De Filippis, N; De Palma, M; De Robertis, G; Donvito, G; Fedele, F; Fiore, L; Franco, M; Iaselli, G; Lacalamita, N; Loddo, F; Lusito, L; Maggi, G; Maggi, M; Manna, N; Marangelli, B; My, S; Natali, S; Nuzzo, S; Papagni, G; Piccolomo, S; Pierro, G A; Pinto, C; Pompili, A; Pugliese, G; Rajan, R; Ranieri, A; Romano, F; Roselli, G; Selvaggi, G; Shinde, Y; Silvestris, L; Tupputi, S; Zito, G; Abbiendi, G; Bacchi, W; Benvenuti, A C; Boldini, M; Bonacorsi, D; Braibant-Giacomelli, S; Cafaro, V D; Caiazza, S S; Capiluppi, P; Castro, A; Cavallo, F R; Codispoti, G; Cuffiani, M; D'Antone, I; Dallavalle, G M; Fabbri, F; Fanfani, A; Fasanella, D; Giacomelli, P; Giordano, V; Giunta, M; Grandi, C; Guerzoni, M; Marcellini, S; Masetti, G; Montanari, A; Navarria, F L; Odorici, F; Pellegrini, G; Perrotta, A; Rossi, A M; Rovelli, T; Siroli, G; Torromeo, G; Travaglini, R; Albergo, S; Costa, S; Potenza, R; Tricomi, A; Tuve, C; Barbagli, G; Broccolo, G; Ciulli, V; Civinini, C; D'Alessandro, R; Focardi, E; Frosali, S; Gallo, E; Genta, C; Landi, G; Lenzi, P; Meschini, M; Paoletti, S; Sguazzoni, G; Tropiano, A; Benussi, L; Bertani, M; Bianco, S; Colafranceschi, S; Colonna, D; Fabbri, F; Giardoni, M; Passamonti, L; Piccolo, D; Pierluigi, D; Ponzio, B; Russo, A; Fabbricatore, P; Musenich, R; Benaglia, A; Calloni, M; Cerati, G B; D'Angelo, P; De Guio, F; Farina, F M; Ghezzi, A; Govoni, P; Malberti, M; Malvezzi, S; Martelli, A; Menasce, D; Miccio, V; Moroni, L; Negri, P; Paganoni, M; Pedrini, D; Pullia, A; Ragazzi, S; Redaelli, N; Sala, S; Salerno, R; Tabarelli de Fatis, T; Tancini, V; Taroni, S; Buontempo, S; Cavallo, N; Cimmino, A; De Gruttola, M; Fabozzi, F; Iorio, A O M; Lista, L; Lomidze, D; Noli, P; Paolucci, P; Sciacca, C; Azzi, P; Bacchetta, N; Barcellan, L; Bellan, P; Bellato, M; Benettoni, M; Biasotto, M; Bisello, D; Borsato, E; Branca, A; Carlin, R; Castellani, L; Checchia, P; Conti, E; Dal Corso, F; De Mattia, M; Dorigo, T; Dosselli, U; Fanzago, F; Gasparini, F; Gasparini, U; Giubilato, P; Gonella, F; Gresele, A; Gulmini, M; Kaminskiy, A; Lacaprara, S; Lazzizzera, I; Margoni, M; Maron, G; Mattiazzo, S; Mazzucato, M; Meneghelli, M; Meneguzzo, A T; Michelotto, M; Montecassiano, F; Nespolo, M; Passaseo, M; Pegoraro, M; Perrozzi, L; Pozzobon, N; Ronchese, P; Simonetto, F; Toniolo, N; Torassa, E; Tosi, M; Triossi, A; Vanini, S; Ventura, S; Zotto, P; Zumerle, G; Baesso, P; Berzano, U; Bricola, S; Necchi, M M; Pagano, D; Ratti, S P; Riccardi, C; Torre, P; Vicini, A; Vitulo, P; Viviani, C; Aisa, D; Aisa, S; Babucci, E; Biasini, M; Bilei, G M; Caponeri, B; Checcucci, B; Dinu, N; Fanò, L; Farnesini, L; Lariccia, P; Lucaroni, A; Mantovani, G; Nappi, A; Piluso, A; Postolache, V; Santocchia, A; Servoli, L; Tonoiu, D; Vedaee, A; Volpe, R; Azzurri, P; Bagliesi, G; Bernardini, J; Berretta, L; Boccali, T; Bocci, A; Borrello, L; Bosi, F; Calzolari, F; Castaldi, R; Dell'Orso, R; Fiori, F; Foà, L; Gennai, S; Giassi, A; Kraan, A; Ligabue, F; Lomtadze, T; Mariani, F; Martini, L; Massa, M; Messineo, A; Moggi, A; Palla, F; Palmonari, F; Petragnani, G; Petrucciani, G; Raffaelli, F; Sarkar, S; Segneri, G; Serban, A T; Spagnolo, P; Tenchini, R; Tolaini, S; Tonelli, G; Venturi, A; Verdini, P G; Baccaro, S; Barone, L; Bartoloni, A; Cavallari, F; Dafinei, I; Del Re, D; Di Marco, E; Diemoz, M; Franci, D; Longo, E; Organtini, G; Palma, A; Pandolfi, F; Paramatti, R; Pellegrino, F; Rahatlou, S; Rovelli, C; Alampi, G; Amapane, N; Arcidiacono, R; Argiro, S; Arneodo, M; Biino, C; Borgia, M A; Botta, C; Cartiglia, N; Castello, R; Cerminara, G; Costa, M; Dattola, D; Dellacasa, G; Demaria, N; Dughera, G; Dumitrache, F; Graziano, A; Mariotti, C; Marone, M; Maselli, S; Migliore, E; Mila, G; Monaco, V; Musich, M; Nervo, M; Obertino, M M; Oggero, S; Panero, R; Pastrone, N; Pelliccioni, M; Romero, A; Ruspa, M; Sacchi, R; Solano, A; Staiano, A; Trapani, P P; Trocino, D; Vilela Pereira, A; Visca, L; Zampieri, A; Ambroglini, F; Belforte, S; Cossutti, F; Della Ricca, G; Gobbo, B; Penzo, A; Chang, S; Chung, J; Kim, D H; Kim, G N; Kong, D J; Park, H; Son, D C; Bahk, S Y; Song, S; Jung, S Y; Hong, B; Kim, H; Kim, J H; Lee, K S; Moon, D H; Park, S K; Rhee, H B; Sim, K S; Kim, J; Choi, M; Hahn, G; Park, I C; Choi, S; Choi, Y; Goh, J; Jeong, H; Kim, T J; Lee, J; Lee, S; Janulis, M; Martisiute, D; Petrov, P; Sabonis, T; Castilla Valdez, H; Sánchez Hernández, A; Carrillo Moreno, S; Morelos Pineda, A; Allfrey, P; Gray, R N C; Krofcheck, D; Bernardino Rodrigues, N; Butler, P H; Signal, T; Williams, J C; Ahmad, M; Ahmed, I; Ahmed, W; Asghar, M I; Awan, M I M; Hoorani, H R; Hussain, I; Khan, W A; Khurshid, T; Muhammad, S; Qazi, S; Shahzad, H; Cwiok, M; Dabrowski, R; Dominik, W; Doroba, K; Konecki, M; Krolikowski, J; Pozniak, K; Romaniuk, Ryszard; Zabolotny, W; Zych, P; Frueboes, T; Gokieli, R; Goscilo, L; Górski, M; Kazana, M; Nawrocki, K; Szleper, M; Wrochna, G; Zalewski, P; Almeida, N; Antunes Pedro, L; Bargassa, P; David, A; Faccioli, P; Ferreira Parracho, P G; Freitas Ferreira, M; Gallinaro, M; Guerra Jordao, M; Martins, P; Mini, G; Musella, P; Pela, J; Raposo, L; Ribeiro, P Q; Sampaio, S; Seixas, J; Silva, J; Silva, P; Soares, D; Sousa, M; Varela, J; Wöhri, H K; Altsybeev, I; Belotelov, I; Bunin, P; Ershov, Y; Filozova, I; Finger, M; Finger, M Jr; Golunov, A; Golutvin, I; Gorbounov, N; Kalagin, V; Kamenev, A; Karjavin, V; Konoplyanikov, V; Korenkov, V; Kozlov, G; Kurenkov, A; Lanev, A; Makankin, A; Mitsyn, V V; Moisenz, P; Nikonov, E; Oleynik, D; Palichik, V; Perelygin, V; Petrosyan, A; Semenov, R; Shmatov, S; Smirnov, V; Smolin, D; Tikhonenko, E; Vasil'ev, S; Vishnevskiy, A; Volodko, A; Zarubin, A; Zhiltsov, V; Bondar, N; Chtchipounov, L; Denisov, A; Gavrikov, Y; Gavrilov, G; Golovtsov, V; Ivanov, Y; Kim, V; Kozlov, V; Levchenko, P; Obrant, G; Orishchin, E; Petrunin, A; Shcheglov, Y; Shchetkovskiy, A; Sknar, V; Smirnov, I; Sulimov, V; Tarakanov, V; Uvarov, L; Vavilov, S; Velichko, G; Volkov, S; Vorobyev, A; Andreev, Yu; Anisimov, A; Antipov, P; Dermenev, A; Gninenko, S; Golubev, N; Kirsanov, M; Krasnikov, N; Matveev, V; Pashenkov, A; Postoev, V E; Solovey, A; Solovey, A; Toropin, A; Troitsky, S; Baud, A; Epshteyn, V; Gavrilov, V; Ilina, N; Kaftanov, V; Kolosov, V; Kossov, M; Krokhotin, A; Kuleshov, S; Oulianov, A; Safronov, G; Semenov, S; Shreyber, I; Stolin, V; Vlasov, E; Zhokin, A; Boos, E; Dubinin, M; Dudko, L; Ershov, A; Gribushin, A; Klyukhin, V; Kodolova, O; Lokhtin, I; Petrushanko, S; Sarycheva, L; Savrin, V; Snigirev, A; Vardanyan, I; Dremin, I; Kirakosyan, M; Konovalova, N; Rusakov, S V; Vinogradov, A; Akimenko, S; Artamonov, A; Azhgirey, I; Bitioukov, S; Burtovoy, V; Grishin, V; Kachanov, V; Konstantinov, D; Krychkine, V; Levine, A; Lobov, I; Lukanin, V; Mel'nik, Y; Petrov, V; Ryutin, R; Slabospitsky, S; Sobol, A; Sytine, A; Tourtchanovitch, L; Troshin, S; Tyurin, N; Uzunian, A; Volkov, A; Adzic, P; Djordjevic, M; Jovanovic, D; Krpic, D; Maletic, D; Puzovic, J; Smiljkovic, N; Aguilar-Benitez, M; Alberdi, J; Alcaraz Maestre, J; Arce, P; Barcala, J M; Battilana, C; Burgos Lazaro, C; Caballero Bejar, J; Calvo, E; Cardenas Montes, M; Cepeda, M; Cerrada, M; Chamizo Llatas, M; Clemente, F; Colino, N; Daniel, M; De La Cruz, B; Delgado Peris, A; Diez Pardos, C; Fernandez Bedoya, C; Fernández Ramos, J P; Ferrando, A; Flix, J; Fouz, M C; Garcia-Abia, P; Garcia-Bonilla, A C; Gonzalez Lopez, O; Goy Lopez, S; Hernandez, J M; Josa, M I; Marin, J; Merino, G; Molina, J; Molinero, A; Navarrete, J J; Oller, J C; Puerta Pelayo, J; Romero, L; Santaolalla, J; Villanueva Munoz, C; Willmott, C; Yuste, C; Albajar, C; Blanco Otano, M; de Trocóniz, J F; Garcia Raboso, A; Lopez Berengueres, J O; Cuevas, J; Fernandez Menendez, J; Gonzalez Caballero, I; Lloret Iglesias, L; Naves Sordo, H; Vizan Garcia, J M; Cabrillo, I J; Calderon, A; Chuang, S H; Diaz Merino, I; Diez Gonzalez, C; Duarte Campderros, J; Fernandez, M; Gomez, G; Gonzalez Sanchez, J; Gonzalez Suarez, R; Jorda, C; Lobelle Pardo, P; Lopez Virto, A; Marco, J; Marco, R; Martinez Rivero, C; Martinez Ruiz del Arbol, P; Matorras, F; Rodrigo, T; Ruiz Jimeno, A; Scodellaro, L; Sobron Sanudo, M; Vila, I; Vilar Cortabitarte, R; Abbaneo, D; Albert, E; Alidra, M; Ashby, S; Auffray, E; Baechler, J; Baillon, P; Ball, A H; Bally, S L; Barney, D; Beaudette, F; Bellan, R; Benedetti, D; Benelli, G; Bernet, C; Bloch, P; Bolognesi, S; Bona, M; Bos, J; Bourgeois, N; Bourrel, T; Breuker, H; Bunkowski, K; Campi, D; Camporesi, T; Cano, E; Cattai, A; Chatelain, J P; Chauvey, M; Christiansen, T; Coarasa Perez, J A; Conde Garcia, A; Covarelli, R; Curé, B; De Roeck, A; Delachenal, V; Deyrail, D; Di Vincenzo, S; Dos Santos, S; Dupont, T; Edera, L M; Elliott-Peisert, A; Eppard, M; Favre, M; Frank, N; Funk, W; Gaddi, A; Gastal, M; Gateau, M; Gerwig, H; Gigi, D; Gill, K; Giordano, D; Girod, J P; Glege, F; Gomez-Reino Garrido, R; Goudard, R; Gowdy, S; Guida, R; Guiducci, L; Gutleber, J; Hansen, M; Hartl, C; Harvey, J; Hegner, B; Hoffmann, H F; Holzner, A; Honma, A; Huhtinen, M; Innocente, V; Janot, P; Le Godec, G; Lecoq, P; Leonidopoulos, C; Loos, R; Lourenço, C; Lyonnet, A; Macpherson, A; Magini, N; Maillefaud, J D; Maire, G; Mäki, T; Malgeri, L; Mannelli, M; Masetti, L; Meijers, F; Meridiani, P; Mersi, S; Meschi, E; Meynet Cordonnier, A; Moser, R; Mulders, M; Mulon, J; Noy, M; Oh, A; Olesen, G; Onnela, A; Orimoto, T; Orsini, L; Perez, E; Perinic, G; Pernot, J F; Petagna, P; Petiot, P; Petrilli, A; Pfeiffer, A; Pierini, M; Pimiä, M; Pintus, R; Pirollet, B; Postema, H; Racz, A; Ravat, S; Rew, S B; Rodrigues Antunes, J; Rolandi, G; Rovere, M; Ryjov, V; Sakulin, H; Samyn, D; Sauce, H; Schäfer, C; Schlatter, W D; Schröder, M; Schwick, C; Sciaba, A; Segoni, I; Sharma, A; Siegrist, N; Siegrist, P; Sinanis, N; Sobrier, T; Sphicas, P; Spiga, D; Spiropulu, M; Stöckli, F; Traczyk, P; Tropea, P; Troska, J; Tsirou, A; Veillet, L; Veres, G I; Voutilainen, M; Wertelaers, P; Zanetti, M; Bertl, W; Deiters, K; Erdmann, W; Gabathuler, K; Horisberger, R; Ingram, Q; Kaestli, H C; König, S; Kotlinski, D; Langenegger, U; Meier, F; Renker, D; Rohe, T; Sibille, J; Starodumov, A; Betev, B; Caminada, L; Chen, Z; Cittolin, S; Da Silva Di Calafiori, D R; Dambach, S; Dissertori, G; Dittmar, M; Eggel, C; Eugster, J; Faber, G; Freudenreich, K; Grab, C; Hervé, A; Hintz, W; Lecomte, P; Luckey, P D; Lustermann, W; Marchica, C; Milenovic, P; Moortgat, F; Nardulli, A; Nessi-Tedaldi, F; Pape, L; Pauss, F; Punz, T; Rizzi, A; Ronga, F J; Sala, L; Sanchez, A K; Sawley, M C; Sordini, V; Stieger, B; Tauscher, L; Thea, A; Theofilatos, K; Treille, D; Trüb, P; Weber, M; Wehrli, L; Weng, J; Zelepoukine, S; Amsler, C; Chiochia, V; De Visscher, S; Regenfus, C; Robmann, P; Rommerskirchen, T; Schmidt, A; Tsirigkas, D; Wilke, L; Chang, Y H; Chen, E A; Chen, W T; Go, A; Kuo, C M; Li, S W; Lin, W; Bartalini, P; Chang, P; Chao, Y; Chen, K F; Hou, W S; Hsiung, Y; Lei, Y J; Lin, S W; Lu, R S; Schümann, J; Shiu, J G; Tzeng, Y M; Ueno, K; Velikzhanin, Y; Wang, C C; Wang, M; Adiguzel, A; Ayhan, A; Azman Gokce, A; Bakirci, M N; Cerci, S; Dumanoglu, I; Eskut, E; Girgis, S; Gurpinar, E; Hos, I; Karaman, T; Karaman, T; Kayis Topaksu, A; Kurt, P; Önengüt, G; Önengüt Gökbulut, G; Ozdemir, K; Ozturk, S; Polatöz, A; Sogut, K; Tali, B; Topakli, H; Uzun, D; Vergili, L N; Vergili, M; Akin, I V; Aliev, T; Bilmis, S; Deniz, M; Gamsizkan, H; Guler, A M; Öcalan, K; Serin, M; Sever, R; Surat, U E; Zeyrek, M; Deliomeroglu, M; Demir, D; Gülmez, E; Halu, A; Isildak, B; Kaya, M; Kaya, O; Ozkorucuklu, S; Sonmez, N; Levchuk, L; Lukyanenko, S; Soroka, D; Zub, S; Bostock, F; Brooke, J J; Cheng, T L; Cussans, D; Frazier, R; Goldstein, J; Grant, N; Hansen, M; Heath, G P; Heath, H F; Hill, C; Huckvale, B; Jackson, J; Mackay, C K; Metson, S; Newbold, D M; Nirunpong, K; Smith, V J; Velthuis, J; Walton, R; Bell, K W; Brew, C; Brown, R M; Camanzi, B; Cockerill, D J A; Coughlan, J A; Geddes, N I; Harder, K; Harper, S; Kennedy, B W; Murray, P; Shepherd-Themistocleous, C H; Tomalin, I R; Williams, J H; Womersley, W J; Worm, S D; Bainbridge, R; Ball, G; Ballin, J; Beuselinck, R; Buchmuller, O; Colling, D; Cripps, N; Davies, G; Della Negra, M; Foudas, C; Fulcher, J; Futyan, D; Hall, G; Hays, J; Iles, G; Karapostoli, G; MacEvoy, B C; Magnan, A M; Marrouche, J; Nash, J; Nikitenko, A; Papageorgiou, A; Pesaresi, M; Petridis, K; Pioppi, M; Raymond, D M; Rompotis, N; Rose, A; Ryan, M J; Seez, C; Sharp, P; Sidiropoulos, G; Stettler, M; Stoye, M; Takahashi, M; Tapper, A; Timlin, C; Tourneur, S; Vazquez Acosta, M; Virdee, T; Wakefield, S; Wardrope, D; Whyntie, T; Wingham, M; Cole, J E; Goitom, I; Hobson, P R; Khan, A; Kyberd, P; Leslie, D; Munro, C; Reid, I D; Siamitros, C; Taylor, R; Teodorescu, L; Yaselli, I; Bose, T; Carleton, M; Hazen, E; Heering, A H; Heister, A; John, J St; Lawson, P; Lazic, D; Osborne, D; Rohlf, J; Sulak, L; Wu, S; Andrea, J; Avetisyan, A; Bhattacharya, S; Chou, J P; Cutts, D; Esen, S; Kukartsev, G; Landsberg, G; Narain, M; Nguyen, D; Speer, T; Tsang, K V; Breedon, R; Calderon De La Barca Sanchez, M; Case, M; Cebra, D; Chertok, M; Conway, J; Cox, P T; Dolen, J; Erbacher, R; Friis, E; Ko, W; Kopecky, A; Lander, R; Lister, A; Liu, H; Maruyama, S; Miceli, T; Nikolic, M; Pellett, D; Robles, J; Searle, M; Smith, J; Squires, M; Stilley, J; Tripathi, M; Vasquez Sierra, R; Veelken, C; Andreev, V; Arisaka, K; Cline, D; Cousins, R; Erhan, S; Hauser, J; Ignatenko, M; Jarvis, C; Mumford, J; Plager, C; Rakness, G; Schlein, P; Tucker, J; Valuev, V; Wallny, R; Yang, X; Babb, J; Bose, M; Chandra, A; Clare, R; Ellison, J A; Gary, J W; Hanson, G; Jeng, G Y; Kao, S C; Liu, F; Liu, H; Luthra, A; Nguyen, H; Pasztor, G; Satpathy, A; Shen, B C; Stringer, R; Sturdy, J; Sytnik, V; Wilken, R; Wimpenny, S; Branson, J G; Dusinberre, E; Evans, D; Golf, F; Kelley, R; Lebourgeois, M; Letts, J; Lipeles, E; Mangano, B; Muelmenstaedt, J; Norman, M; Padhi, S; Petrucci, A; Pi, H; Pieri, M; Ranieri, R; Sani, M; Sharma, V; Simon, S; Würthwein, F; Yagil, A; Campagnari, C; D'Alfonso, M; Danielson, T; Garberson, J; Incandela, J; Justus, C; Kalavase, P; Koay, S A; Kovalskyi, D; Krutelyov, V; Lamb, J; Lowette, S; Pavlunin, V; Rebassoo, F; Ribnik, J; Richman, J; Rossin, R; Stuart, D; To, W; Vlimant, J R; Witherell, M; Apresyan, A; Bornheim, A; Bunn, J; Chiorboli, M; Gataullin, M; Kcira, D; Litvine, V; Ma, Y; Newman, H B; Rogan, C; Timciuc, V; Veverka, J; Wilkinson, R; Yang, Y; Zhang, L; Zhu, K; Zhu, R Y; Akgun, B; Carroll, R; Ferguson, T; Jang, D W; Jun, S Y; Paulini, M; Russ, J; Terentyev, N; Vogel, H; Vorobiev, I; Cumalat, J P; Dinardo, M E; Drell, B R; Ford, W T; Heyburn, B; Luiggi Lopez, E; Nauenberg, U; Stenson, K; Ulmer, K; Wagner, S R; Zang, S L; Agostino, L; Alexander, J; Blekman, F; Cassel, D; Chatterjee, A; Das, S; Gibbons, L K; Heltsley, B; Hopkins, W; Khukhunaishvili, A; Kreis, B; Kuznetsov, V; Patterson, J R; Puigh, D; Ryd, A; Shi, X; Stroiney, S; Sun, W; Teo, W D; Thom, J; Vaughan, J; Weng, Y; Wittich, P; Beetz, C P; Cirino, G; Sanzeni, C; Winn, D; Abdullin, S; Afaq, M A; Albrow, M; Ananthan, B; Apollinari, G; Atac, M; Badgett, W; Bagby, L; Bakken, J A; Baldin, B; Banerjee, S; Banicz, K; Bauerdick, L A T; Beretvas, A; Berryhill, J; Bhat, P C; Biery, K; Binkley, M; Bloch, I; Borcherding, F; Brett, A M; Burkett, K; Butler, J N; Chetluru, V; Cheung, H W K; Chlebana, F; Churin, I; Cihangir, S; Crawford, M; Dagenhart, W; Demarteau, M; Derylo, G; Dykstra, D; Eartly, D P; Elias, J E; Elvira, V D; Evans, D; Feng, L; Fischler, M; Fisk, I; Foulkes, S; Freeman, J; Gartung, P; Gottschalk, E; Grassi, T; Green, D; Guo, Y; Gutsche, O; Hahn, A; Hanlon, J; Harris, R M; Holzman, B; Howell, J; Hufnagel, D; James, E; Jensen, H; Johnson, M; Jones, C D; Joshi, U; Juska, E; Kaiser, J; Klima, B; Kossiakov, S; Kousouris, K; Kwan, S; Lei, C M; Limon, P; Lopez Perez, J A; Los, S; Lueking, L; Lukhanin, G; Lusin, S; Lykken, J; Maeshima, K; Marraffino, J M; Mason, D; McBride, P; Miao, T; Mishra, K; Moccia, S; Mommsen, R; Mrenna, S; Muhammad, A S; Newman-Holmes, C; Noeding, C; O'Dell, V; Prokofyev, O; Rivera, R; Rivetta, C H; Ronzhin, A; Rossman, P; Ryu, S; Sekhri, V; Sexton-Kennedy, E; Sfiligoi, I; Sharma, S; Shaw, T M; Shpakov, D; Skup, E; Smith, R P; Soha, A; Spalding, W J; Spiegel, L; Suzuki, I; Tan, P; Tanenbaum, W; Tkaczyk, S; Trentadue, R; Uplegger, L; Vaandering, E W; Vidal, R; Whitmore, J; Wicklund, E; Wu, W; Yarba, J; Yumiceva, F; Yun, J C; Acosta, D; Avery, P; Barashko, V; Bourilkov, D; Chen, M; Di Giovanni, G P; Dobur, D; Drozdetskiy, A; Field, R D; Fu, Y; Furic, I K; Gartner, J; Holmes, D; Kim, B; Klimenko, S; Konigsberg, J; Korytov, A; Kotov, K; Kropivnitskaya, A; Kypreos, T; Madorsky, A; Matchev, K; Mitselmakher, G; Pakhotin, Y; Piedra Gomez, J; Prescott, C; Rapsevicius, V; Remington, R; Schmitt, M; Scurlock, B; Wang, D; Yelton, J; Ceron, C; Gaultney, V; Kramer, L; Lebolo, L M; Linn, S; Markowitz, P; Martinez, G; Rodriguez, J L; Adams, T; Askew, A; Baer, H; Bertoldi, M; Chen, J; Dharmaratna, W G D; Gleyzer, S V; Haas, J; Hagopian, S; Hagopian, V; Jenkins, M; Johnson, K F; Prettner, E; Prosper, H; Sekmen, S; Baarmand, M M; Guragain, S; Hohlmann, M; Kalakhety, H; Mermerkaya, H; Ralich, R; Vodopiyanov, I; Abelev, B; Adams, M R; Anghel, I M; Apanasevich, L; Bazterra, V E; Betts, R R; Callner, J; Castro, M A; Cavanaugh, R; Dragoiu, C; Garcia-Solis, E J; Gerber, C E; Hofman, D J; Khalatian, S; Mironov, C; Shabalina, E; Smoron, A; Varelas, N; Akgun, U; Albayrak, E A; Ayan, A S; Bilki, B; Briggs, R; Cankocak, K; Chung, K; Clarida, W; Debbins, P; Duru, F; Ingram, F D; Lae, C K; McCliment, E; Merlo, J P; Mestvirishvili, A; Miller, M J; Moeller, A; Nachtman, J; Newsom, C R; Norbeck, E; Olson, J; Onel, Y; Ozok, F; Parsons, J; Schmidt, I; Sen, S; Wetzel, J; Yetkin, T; Yi, K; Barnett, B A; Blumenfeld, B; Bonato, A; Chien, C Y; Fehling, D; Giurgiu, G; Gritsan, A V; Guo, Z J; Maksimovic, P; Rappoccio, S; Swartz, M; Tran, N V; Zhang, Y; Baringer, P; Bean, A; Grachov, O; Murray, M; Radicci, V; Sanders, S; Wood, J S; Zhukova, V; Bandurin, D; Bolton, T; Kaadze, K; Liu, A; Maravin, Y; Onoprienko, D; Svintradze, I; Wan, Z; Gronberg, J; Hollar, J; Lange, D; Wright, D; Baden, D; Bard, R; Boutemeur, M; Eno, S C; Ferencek, D; Hadley, N J; Kellogg, R G; Kirn, M; Kunori, S; Rossato, K; Rumerio, P; Santanastasio, F; Skuja, A; Temple, J; Tonjes, M B; Tonwar, S C; Toole, T; Twedt, E; Alver, B; Bauer, G; Bendavid, J; Busza, W; Butz, E; Cali, I A; Chan, M; D'Enterria, D; Everaerts, P; Gomez Ceballos, G; Hahn, K A; Harris, P; Jaditz, S; Kim, Y; Klute, M; Lee, Y J; Li, W; Loizides, C; Ma, T; Miller, M; Nahn, S; Paus, C; Roland, C; Roland, G; Rudolph, M; Stephans, G; Sumorok, K; Sung, K; Vaurynovich, S; Wenger, E A; Wyslouch, B; Xie, S; Yilmaz, Y; Yoon, A S; Bailleux, D; Cooper, S I; Cushman, P; Dahmes, B; De Benedetti, A; Dolgopolov, A; Dudero, P R; Egeland, R; Franzoni, G; Haupt, J; Inyakin, A; Klapoetke, K; Kubota, Y; Mans, J; Mirman, N; Petyt, D; Rekovic, V; Rusack, R; Schroeder, M; Singovsky, A; Zhang, J; Cremaldi, L M; Godang, R; Kroeger, R; Perera, L; Rahmat, R; Sanders, D A; Sonnek, P; Summers, D; Bloom, K; Bockelman, B; Bose, S; Butt, J; Claes, D R; Dominguez, A; Eads, M; Keller, J; Kelly, T; Kravchenko, I; Lazo-Flores, J; Lundstedt, C; Malbouisson, H; Malik, S; Snow, G R; Baur, U; Iashvili, I; Kharchilava, A; Kumar, A; Smith, K; Strang, M; Alverson, G; Barberis, E; Boeriu, O; Eulisse, G; Govi, G; McCauley, T; Musienko, Y; Muzaffar, S; Osborne, I; Paul, T; Reucroft, S; Swain, J; Taylor, L; Tuura, L; Anastassov, A; Gobbi, B; Kubik, A; Ofierzynski, R A; Pozdnyakov, A; Schmitt, M; Stoynev, S; Velasco, M; Won, S; Antonelli, L; Berry, D; Hildreth, M; Jessop, C; Karmgard, D J; Kolberg, T; Lannon, K; Lynch, S; Marinelli, N; Morse, D M; Ruchti, R; Slaunwhite, J; Warchol, J; Wayne, M; Bylsma, B; Durkin, L S; Gilmore, J; Gu, J; Killewald, P; Ling, T Y; Williams, G; Adam, N; Berry, E; Elmer, P; Garmash, A; Gerbaudo, D; Halyo, V; Hunt, A; Jones, J; Laird, E; Marlow, D; Medvedeva, T; Mooney, M; Olsen, J; Piroué, P; Stickland, D; Tully, C; Werner, J S; Wildish, T; Xie, Z; Zuranski, A; Acosta, J G; Bonnett Del Alamo, M; Huang, X T; Lopez, A; Mendez, H; Oliveros, S; Ramirez Vargas, J E; Santacruz, N; Zatzerklyany, A; Alagoz, E; Antillon, E; Barnes, V E; Bolla, G; Bortoletto, D; Everett, A; Garfinkel, A F; Gecse, Z; Gutay, L; Ippolito, N; Jones, M; Koybasi, O; Laasanen, A T; Leonardo, N; Liu, C; Maroussov, V; Merkel, P; Miller, D H; Neumeister, N; Sedov, A; Shipsey, I; Yoo, H D; Zheng, Y; Jindal, P; Parashar, N; Cuplov, V; Ecklund, K M; Geurts, F J M; Liu, J H; Maronde, D; Matveev, M; Padley, B P; Redjimi, R; Roberts, J; Sabbatini, L; Tumanov, A; Betchart, B; Bodek, A; Budd, H; Chung, Y S; de Barbaro, P; Demina, R; Flacher, H; Gotra, Y; Harel, A; Korjenevski, S; Miner, D C; Orbaker, D; Petrillo, G; Vishnevskiy, D; Zielinski, M; Bhatti, A; Demortier, L; Goulianos, K; Hatakeyama, K; Lungu, G; Mesropian, C; Yan, M; Atramentov, O; Bartz, E; Gershtein, Y; Halkiadakis, E; Hits, D; Lath, A; Rose, K; Schnetzer, S; Somalwar, S; Stone, R; Thomas, S; Watts, T L; Cerizza, G; Hollingsworth, M; Spanier, S; Yang, Z C; York, A; Asaadi, J; Aurisano, A; Eusebi, R; Golyash, A; Gurrola, A; Kamon, T; Nguyen, C N; Pivarski, J; Safonov, A; Sengupta, S; Toback, D; Weinberger, M; Akchurin, N; Berntzon, L; Gumus, K; Jeong, C; Kim, H; Lee, S W; Popescu, S; Roh, Y; Sill, A; Volobouev, I; Washington, E; Wigmans, R; Yazgan, E; Engh, D; Florez, C; Johns, W; Pathak, S; Sheldon, P; Andelin, D; Arenton, M W; Balazs, M; Boutle, S; Buehler, M; Conetti, S; Cox, B; Hirosky, R; Ledovskoy, A; Neu, C; Phillips II, D; Ronquest, M; Yohay, R; Gollapinni, S; Gunthoti, K; Harr, R; Karchin, P E; Mattson, M; Sakharov, A; Anderson, M; Bachtis, M; Bellinger, J N; Carlsmith, D; Crotty, I; Dasu, S; Dutta, S; Efron, J; Feyzi, F; Flood, K; Gray, L; Grogg, K S; Grothe, M; Hall-Wilton, R; Jaworski, M; Klabbers, P; Klukas, J; Lanaro, A; Lazaridis, C; Leonard, J; Loveless, R; Magrans de Abril, M; Mohapatra, A; Ott, G; Polese, G; Reeder, D; Savin, A; Smith, W H; Sourkov, A; Swanson, J; Weinberg, M; Wenman, D; Wensveen, M; White, A

    2010-01-01

    This paper describes the calibration procedure for the drift tubes of the CMS barrel muon system and reports the main results obtained with data collected during a high statistics cosmic ray data-taking period. The main goal of the calibration is to determine, for each drift cell, the minimum time delay for signals relative to the trigger, accounting for the drift velocity within the cell. The accuracy of the calibration procedure is influenced by the random arrival time of cosmic muons. A more refined analysis of the drift velocity was performed during the offline reconstruction phase, which takes into account this feature of cosmic ray events.

  2. Integrated Performance Testing Workshop - Supplemental Materials (Scripts and Procedures)

    Energy Technology Data Exchange (ETDEWEB)

    Baum, Gregory A.

    2014-02-01

    A variety of performance tests are described relating to: Material Transfers; Emergency Evacuation; Alarm Response Assessment; and an Enhanced Limited Scope Performance Test (ELSPT). Procedures are given for: nuclear material physical inventory and discrepancy; material transfers; and emergency evacuation.

  3. A New Approach to the Internal Calibration of Reverberation Mapping Spectra

    CERN Document Server

    Fausnaugh, M M

    2016-01-01

    We present a new procedure for the internal (night-to-night) calibration of time series spectra, with specific applications to optical AGN reverberation mapping data. The traditional calibration technique assumes that the narrow [OIII]$\\lambda$5007 emission line profile is constant in time; given a reference [OIII]$\\lambda$5007 line profile, nightly spectra are aligned by fitting for a wavelength shift, a flux rescaling factor, and a change in the spectroscopic resolution. We propose the following modifications to this procedure: 1) we stipulate a constant spectral resolution for the final calibrated spectra, 2) we employ a more flexible model for changes in the spectral resolution, and 3) we use a Bayesian modeling framework to assess uncertainties in the calibration. In a test case using data for MCG+08-11-011, these modifications result in a calibration precision of $\\sim\\! 1$ millimagnitude, which is approximately a factor of five improvement over the traditional technique. At this level, other systematic...

  4. SIR-B experiments in Japan. I - Sensor calibration experiment

    Science.gov (United States)

    Ichinose, Masaru; Echizenya, Yoshimatsu; Kamata, Mitsuhiro; Kawai, Eiji; Hiromoto, Norihisa

    1988-03-01

    A sensor calibration experiment was proposed as part of SIR-B experiments in Japan, together with the rice crop experiment and the ocean oil-pollution detection experiment. This sensor calibration experiment was intended (1) to establish a transfer function from image data to radar backscattering characteristics, (2) to evaluate 3-dB resolutions, (3) to verify the ability to resolve two closely-spaced targets, and (4) to clarify sidelobe structures due to range and azimuth compressions. The disused Akita Airport was chosen as the main test site for the calibration experiment on the first three objectives. This paper describes the test site, the design of the corner reflectors, and briefly predicts the results.

  5. Quantitative Analysis of Carbon Steel with Multi-Line Internal Standard Calibration Method Using Laser-Induced Breakdown Spectroscopy.

    Science.gov (United States)

    Pan, Congyuan; Du, Xuewei; An, Ning; Zeng, Qiang; Wang, Shengbo; Wang, Qiuping

    2016-04-01

    A multi-line internal standard calibration method is proposed for the quantitative analysis of carbon steel using laser-induced breakdown spectroscopy (LIBS). A procedure based on the method was adopted to select the best calibration curves and the corresponding emission lines pairs automatically. Laser-induced breakdown spectroscopy experiments with carbon steel samples were performed, and C, Cr, and Mn were analyzed via the proposed method. Calibration curves of these elements were constructed via a traditional single line internal standard calibration method and a multi-line internal standard calibration method. The calibration curves obtained were evaluated with the determination coefficient, the root mean square error of cross-validation, and the average relative error of cross-validation. All of the parameters were improved significantly with the proposed method. The results show that accurate and stable calibration curves can be obtained efficiently via the multi-line internal standard calibration method.

  6. Calibration of "Babyline" RP instruments

    CERN Multimedia

    2015-01-01

      If you have old RP instrumentation of the “Babyline” type, as shown in the photo, please contact the Radiation Protection Group (Joffrey Germa, 73171) to have the instrument checked and calibrated. Thank you. Radiation Protection Group

  7. Field calibration of cup anemometers

    Energy Technology Data Exchange (ETDEWEB)

    Kristensen, L.; Jensen, G.; Hansen, A.; Kirkegaard, P.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statistical significance of the calibration expressions. It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometers may take more than one month in order to have wind speeds covering a sufficiently large magnitude range in a wind direction sector where we can be sure that the instruments are exposed to identical, simultaneous wind flows. Another main conclusion is that statistical uncertainty must be carefully evaluated since the individual 10 minute wind-speed averages are not statistically independent. (au)

  8. Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material.

    Science.gov (United States)

    Blirup-Jensen, Søren; Johnson, A Myron; Larsen, Marianne

    2008-01-01

    We present a practical protocol for the assignment of values to serum proteins in a Target Material using a Reference Material. This protocol is based on the model of Direct Value Transfer between serum matrices and is intended to improve the value assignment of commercial calibrators using the Reference Material CRM 470 (now labeled ERM-DA 470) or similar reference materials. The procedure describes the general as well as the practical principles involved in the value assignment (with examples). The practical transfer protocol is based on multiple assays of 6 dilutions of the Reference Material and 6 dilutions of the Target Material. The transfer protocol requires several measurements a day repeated on several days, an important prerequisite being that all reconstitutions and dilutions are controlled by weighing thus reducing uncertainty in the transfer. In open systems that allow the use of the Reference Material as calibrator and the Target Material as samples, the proportionality of the two materials (the presence or absence of matrix effects) can now be directly assessed by evaluating a single regression plot. If no matrix effects are found, the regression line will pass through zero with a slope equal to the ratio of the concentrations of the two materials. In closed systems, the dedicated commercial calibrator has to be used as such; the Reference Material and the Target Material are now assayed as samples against this calibrator. Two regression plots are therefore obtained; if no matrix effects are present among the two materials and the calibrator, both the Reference and Target Materials will show zero intercepts, and the ratio of the two slopes will equal the ratio of the concentrations.

  9. Infrasound Sensor Calibration and Response

    Science.gov (United States)

    2012-09-01

    functions with faster rise times. SUMMARY We have documented past work on the determination of the calibration constant of the LANL infrasound sensor...Monitoring Technologies 735 Report Documentation Page Form ApprovedOMB No. 0704-0188 Public reporting burden for the collection of information is estimated...National Laboratory ( LANL ) has operated an infrasound sensor calibration chamber that operates over a frequency range of 0.02 to 4 Hz. This chamber has

  10. Bayesian Calibration of Microsimulation Models.

    Science.gov (United States)

    Rutter, Carolyn M; Miglioretti, Diana L; Savarino, James E

    2009-12-01

    Microsimulation models that describe disease processes synthesize information from multiple sources and can be used to estimate the effects of screening and treatment on cancer incidence and mortality at a population level. These models are characterized by simulation of individual event histories for an idealized population of interest. Microsimulation models are complex and invariably include parameters that are not well informed by existing data. Therefore, a key component of model development is the choice of parameter values. Microsimulation model parameter values are selected to reproduce expected or known results though the process of model calibration. Calibration may be done by perturbing model parameters one at a time or by using a search algorithm. As an alternative, we propose a Bayesian method to calibrate microsimulation models that uses Markov chain Monte Carlo. We show that this approach converges to the target distribution and use a simulation study to demonstrate its finite-sample performance. Although computationally intensive, this approach has several advantages over previously proposed methods, including the use of statistical criteria to select parameter values, simultaneous calibration of multiple parameters to multiple data sources, incorporation of information via prior distributions, description of parameter identifiability, and the ability to obtain interval estimates of model parameters. We develop a microsimulation model for colorectal cancer and use our proposed method to calibrate model parameters. The microsimulation model provides a good fit to the calibration data. We find evidence that some parameters are identified primarily through prior distributions. Our results underscore the need to incorporate multiple sources of variability (i.e., due to calibration data, unknown parameters, and estimated parameters and predicted values) when calibrating and applying microsimulation models.

  11. Pressures Detector Calibration and Measurement

    CERN Document Server

    AUTHOR|(CDS)2156315

    2016-01-01

    This is report of my first and second projects (of 3) in NA61. I did data taking and analysis in order to do calibration of pressure detectors and verified it. I analyzed the data by ROOT software using the C ++ programming language. The first part of my project was determination of calibration factor of pressure sensors. Based on that result, I examined the relation between pressure drop, gas flow rate of in paper filter and its diameter.

  12. Beam Imaging and Luminosity Calibration

    CERN Document Server

    Klute, Markus; Salfeld-Nebgen, Jakob

    2016-01-01

    We discuss a method to reconstruct two-dimensional proton bunch densities using vertex distributions accumulated during LHC beam-beam scans. The $x$-$y$ correlations in the beam shapes are studied and an alternative luminosity calibration technique is introduced. We demonstrate the method on simulated beam-beam scans and estimate the uncertainty on the luminosity calibration associated to the beam-shape reconstruction to be below 1\\%.

  13. Biofluid Flow Simulations of Embryo Transfer

    Science.gov (United States)

    Shi, W. P.; Ding, D. L.

    2011-09-01

    The investigation of the fluid flow for embryo transfer (ET) procedure may find the way to increase the success rate of the assisted reproductive technologies. In this paper, the transferred liquid flow in the uterine cavity during ET procedure is simulated by a two dimensional multiphase flow model, and the discrete phase model is adopted to trace the embryo motion. Through the investigation on the transferred liquid outline and the track of each embryo in ET cases with different parameters, we summarize the effect of transferred liquid viscosity and distance between catheter tip and uterine fundus. According to the numerical results, we recommend the optimizing standard to perform the ET procedure.

  14. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    Science.gov (United States)

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  15. Camera calibration method of binocular stereo vision based on OpenCV

    Science.gov (United States)

    Zhong, Wanzhen; Dong, Xiaona

    2015-10-01

    Camera calibration, an important part of the binocular stereo vision research, is the essential foundation of 3D reconstruction of the spatial object. In this paper, the camera calibration method based on OpenCV (open source computer vision library) is submitted to make the process better as a result of obtaining higher precision and efficiency. First, the camera model in OpenCV and an algorithm of camera calibration are presented, especially considering the influence of camera lens radial distortion and decentering distortion. Then, camera calibration procedure is designed to compute those parameters of camera and calculate calibration errors. High-accurate profile extraction algorithm and a checkboard with 48 corners have also been used in this part. Finally, results of calibration program are presented, demonstrating the high efficiency and accuracy of the proposed approach. The results can reach the requirement of robot binocular stereo vision.

  16. Development and calibration of buried wire gages for wall shear stress measurements in fluid flow

    Science.gov (United States)

    Murthy, Sreedhara V.; Steinle, Frank W.

    1988-01-01

    Special methods were developed to arrange 'Buried Wire Gage' inserts flush to the contoured flow surfaces of instrument plugs of a boundary-layer flow apparatus. The fabrication process was aimed at producing proper bonding of the sensor wire to the substrate surface, without causing excessive surface waviness. A large number of gages were built and first calibrated for the resistance-temperature characteristics. The gages were then installed in a flow calibration apparatus and operated from a constant temperature anemometer system for a series of flow settings to derive the calibration constants of each of the gages. The flow settings included a range of subsonic freestream Mach numbers in order to help establish the gage calibration characteristics for compressible flow fields. This paper provides a description of the buried wire gage technique, an explanation of the method evolved for making proper gages, the procedure for calibrating the gages and the results of measurements performed for determining the calibration constants.

  17. Calibrating Data from the Hinode/X-Ray Telescope and Associated Uncertainties

    CERN Document Server

    Kobelski, Adam R; Weber, Mark A; McKenzie, David E; Reeves, Katharine K

    2013-01-01

    The X-Ray Telescope (XRT) onboard the Hinode satellite, launched 23 September 2006 by the Japanese Aerospace Exploration Agency (JAXA) is a joint mission between Japan, the United States, and the United Kingdom to study the solar corona. In particular XRT was designed to study solar plasmas with temperatures between 1 and 10 MK with $\\approx1''$ pixels ($\\approx2''$ resolution). Prior to analysis, the data product from this instrument must be properly calibrated and data values quantified in order to assess accurately the information contained within. We present here the standard methods of calibration for these data. The calibration is performed on an empirical basis which uses the least complicated correction that accurately describes the data while suppressing spurious features. By analyzing the uncertainties remaining in the data after calibration, we conclude that the procedure is successful, as the remaining uncertainty after calibration is dominated by photon noise. This calibration software is availab...

  18. Neutron Imaging Calibration to Measure Void Fraction

    Energy Technology Data Exchange (ETDEWEB)

    Geoghegan, Patrick J [ORNL; Bilheux, Hassina Z [ORNL; Sharma, Vishaldeep [ORNL; Fricke, Brian A [ORNL

    2015-01-01

    Void fraction is an intuitive parameter that describes the fraction of vapor in a two-phase flow. It appears as a key variable in most heat transfer and pressure drop correlations used to design evaporating and condensing heat exchangers, as well as determining charge inventory in refrigeration systems. Void fraction measurement is not straightforward, however, and assumptions on the invasiveness of the measuring technique must be made. Neutron radiography or neutron imaging has the potential to be a truly non-invasive void fraction measuring technique but has until recently only offered qualitative descriptions of two-phase flow, in terms of flow maldistributions, for example. This paper describes the calibration approach necessary to employ neutron imaging to measure steady-state void fraction. Experiments were conducted at the High Flux Isotope Reactor (HFIR) Cold Guide 1D neutron imaging facility at Oak Ridge National Laboratory (ORNL), Oak Ridge, TN, USA.

  19. Transfer Pricing

    DEFF Research Database (Denmark)

    Nielsen, Søren Bo

    2014-01-01

    Against a background of rather mixed evidence about transfer pricing practices in multinational enterprises (MNEs) and varying attitudes on the part of tax authorities, this paper explores how multiple aims in transfer pricing can be pursued across four different transfer pricing regimes. A MNE h...

  20. Radiometric calibration status of Landsat-7 and Landsat-5

    Science.gov (United States)

    Barsi, J.A.; Markham, B.L.; Helder, D.L.; Chander, G.

    2007-01-01

    Launched in April 1999, Landsat-7 ETM+ continues to acquire data globally. The Scan Line Corrector in failure in 2003 has affected ground coverage and the recent switch to Bumper Mode operations in April 2007 has degraded the internal geometric accuracy of the data, but the radiometry has been unaffected. The best of the three on-board calibrators for the reflective bands, the Full Aperture Solar Calibrator, has indicated slow changes in the ETM+, but this is believed to be due to contamination on the panel rather then instrument degradation. The Internal Calibrator lamp 2, though it has not been used regularly throughout the whole mission, indicates smaller changes than the FASC since 2003. The changes indicated by lamp 2 are only statistically significant in band 1, circa 0.3% per year, and may be lamp as opposed to instrument degradations. Regular observations of desert targets in the Saharan and Arabian deserts indicate the no change in the ETM+ reflective band response, though the uncertainty is larger and does not preclude the small changes indicated by lamp 2. The thermal band continues to be stable and well-calibrated since an offset error was corrected in late-2000. Launched in 1984, Landsat-5 TM also continues to acquire global data; though without the benefit of an on-board recorder, data can only be acquired where a ground station is within range. Historically, the calibration of the TM reflective bands has used an onboard calibration system with multiple lamps. The calibration procedure for the TM reflective bands was updated in 2003 based on the best estimate at the time, using only one of the three lamps and a cross-calibration with Landsat-7 ETM+. Since then, the Saharan desert sites have been used to validate this calibration model. Problems were found with the lamp based model of up to 13% in band 1. Using the Saharan data, a new model was developed and implemented in the US processing system in April 2007. The TM thermal band was found to have a

  1. A Generic Algorithm for IACT Optical Efficiency Calibration using Muons

    CERN Document Server

    Mitchell, A M W; Parsons, R D

    2015-01-01

    Muons produced in Extensive Air Showers (EAS) generate ring-like images in Imaging Atmospheric Cherenkov Telescopes when travelling near parallel to the optical axis. From geometrical parameters of these images, the absolute amount of light emitted may be calculated analytically. Comparing the amount of light recorded in these images to expectation is a well established technique for telescope optical efficiency calibration. However, this calculation is usually performed under the assumption of an approximately circular telescope mirror. The H.E.S.S. experiment entered its second phase in 2012, with the addition of a fifth telescope with a non-circular 600m$^2$ mirror. Due to the differing mirror shape of this telescope to the original four H.E.S.S. telescopes, adaptations to the standard muon calibration were required. We present a generalised muon calibration procedure, adaptable to telescopes of differing shapes and sizes, and demonstrate its performance on the H.E.S.S. II array.

  2. Improving Photometric Calibration of Meteor Video Camera Systems

    Science.gov (United States)

    Ehlert, Steven; Kingery, Aaron; Cooke, William

    2016-01-01

    Current optical observations of meteors are commonly limited by systematic uncertainties in photometric calibration at the level of approximately 0.5 mag or higher. Future improvements to meteor ablation models, luminous efficiency models, or emission spectra will hinge on new camera systems and techniques that significantly reduce calibration uncertainties and can reliably perform absolute photometric measurements of meteors. In this talk we discuss the algorithms and tests that NASA's Meteoroid Environment Office (MEO) has developed to better calibrate photometric measurements for the existing All-Sky and Wide-Field video camera networks as well as for a newly deployed four-camera system for measuring meteor colors in Johnson-Cousins BV RI filters. In particular we will emphasize how the MEO has been able to address two long-standing concerns with the traditional procedure, discussed in more detail below.

  3. Automatic Chessboard Detection for Intrinsic and Extrinsic Camera Parameter Calibration

    Directory of Open Access Journals (Sweden)

    Jose María Armingol

    2010-03-01

    Full Text Available There are increasing applications that require precise calibration of cameras to perform accurate measurements on objects located within images, and an automatic algorithm would reduce this time consuming calibration procedure. The method proposed in this article uses a pattern similar to that of a chess board, which is found automatically in each image, when no information regarding the number of rows or columns is supplied to aid its detection. This is carried out by means of a combined analysis of two Hough transforms, image corners and invariant properties of the perspective transformation. Comparative analysis with more commonly used algorithms demonstrate the viability of the algorithm proposed, as a valuable tool for camera calibration.

  4. Mathematical Properties of Numerical Inversion for Jet Calibrations

    CERN Document Server

    Cukierman, Aviv

    2016-01-01

    Numerical inversion is a general detector calibration technique that is independent of the underlying spectrum. This procedure is formalized and important statistical properties are presented, using high energy jets at the Large Hadron Collider as an example setting. In particular, numerical inversion is inherently biased and common approximations to the calibrated jet energy tend to over-estimate the resolution. Analytic approximations to the closure and calibrated resolutions are demonstrated to effectively predict the full forms under realistic conditions. Finally, extensions of numerical inversion are presented which can reduce the inherent biases. These methods will be increasingly important to consider with degraded resolution at low jet energies due to a much higher instantaneous luminosity in the near future.

  5. Multi-sensor calibration of low-cost magnetic, angular rate and gravity systems.

    Science.gov (United States)

    Lüken, Markus; Misgeld, Berno J E; Rüschen, Daniel; Leonhardt, Steffen

    2015-10-13

    We present a new calibration procedure for low-cost nine degrees-of-freedom (9DOF) magnetic, angular rate and gravity (MARG) sensor systems, which relies on a calibration cube, a reference table and a body sensor network (BSN). The 9DOF MARG sensor is part of our recently-developed "Integrated Posture and Activity Network by Medit Aachen" (IPANEMA) BSN. The advantage of this new approach is the use of the calibration cube, which allows for easy integration of two sensor nodes of the IPANEMA BSN. One 9DOF MARG sensor node is thereby used for calibration; the second 9DOF MARG sensor node is used for reference measurements. A novel algorithm uses these measurements to further improve the performance of the calibration procedure by processing arbitrarily-executed motions. In addition, the calibration routine can be used in an alignment procedure to minimize errors in the orientation between the 9DOF MARG sensor system and a motion capture inertial reference system. A two-stage experimental study is conducted to underline the performance of our calibration procedure. In both stages of the proposed calibration procedure, the BSN data, as well as reference tracking data are recorded. In the first stage, the mean values of all sensor outputs are determined as the absolute measurement offset to minimize integration errors in the derived movement model of the corresponding body segment. The second stage deals with the dynamic characteristics of the measurement system where the dynamic deviation of the sensor output compared to a reference system is Sensors 2015, 15 25920 corrected. In practical validation experiments, this procedure showed promising results with a maximum RMS error of 3.89°.

  6. Multi-Sensor Calibration of Low-Cost Magnetic, Angular Rate and Gravity Systems

    Directory of Open Access Journals (Sweden)

    Markus Lüken

    2015-10-01

    Full Text Available We present a new calibration procedure for low-cost nine degrees-of-freedom (9DOF magnetic, angular rate and gravity (MARG sensor systems, which relies on a calibration cube, a reference table and a body sensor network (BSN. The 9DOF MARG sensor is part of our recently-developed “Integrated Posture and Activity Network by Medit Aachen” (IPANEMA BSN. The advantage of this new approach is the use of the calibration cube, which allows for easy integration of two sensor nodes of the IPANEMA BSN. One 9DOF MARG sensor node is thereby used for calibration; the second 9DOF MARG sensor node is used for reference measurements. A novel algorithm uses these measurements to further improve the performance of the calibration procedure by processing arbitrarily-executed motions. In addition, the calibration routine can be used in an alignment procedure to minimize errors in the orientation between the 9DOF MARG sensor system and a motion capture inertial reference system. A two-stage experimental study is conducted to underline the performance of our calibration procedure. In both stages of the proposed calibration procedure, the BSN data, as well as reference tracking data are recorded. In the first stage, the mean values of all sensor outputs are determined as the absolute measurement offset to minimize integration errors in the derived movement model of the corresponding body segment. The second stage deals with the dynamic characteristics of the measurement system where the dynamic deviation of the sensor output compared to a reference system is Sensors 2015, 15 25920 corrected. In practical validation experiments, this procedure showed promising results with a maximum RMS error of 3.89°.

  7. Photometric Calibration of the [$\\alpha$/Fe] Element: I. Calibration with UBV Photometry

    CERN Document Server

    Karaali, S; Bilir, S

    2016-01-01

    We present the calibration of the [$\\alpha$/Fe] element in terms of the ultra-violet excess for 469 dwarf stars with $0.325<(B-V)_0 \\leq 0.775$ mag corresponding the spectral type range F0-K2. The star sample is separated into nine sub-samples with equal range in $(B-V)_0$ colour, $\\Delta(B-V)_0=0.05$ mag, and a third degree polynomial is fitted to each dataset. Our calibrations provide [$\\alpha$/Fe] elements in the range [0.0, 0.4]. We applied the procedure to two sets of field stars and two sets of clusters. The mean and the corresponding standard deviation of the residuals for 43 field stars taken from the Hypatia catalogue are [$\\alpha$/Fe]=-0.090 and $\\sigma =0.102$ dex, while for the 39 ones taken from the same catalogue of stars used in the calibration are [$\\alpha$/Fe]=-0.009 and $\\sigma = 0.079$ dex, respectively. We showed that the differences between the mean of the residuals and standard deviations for two sets of clusters ([$\\alpha$/Fe]=0.073 and $\\sigma = 0.91$ dex; [$\\alpha$/Fe]=-0.012 and $...

  8. CALIBRATION OF ONLINE ANALYZERS USING NEURAL NETWORKS

    Energy Technology Data Exchange (ETDEWEB)

    Rajive Ganguli; Daniel E. Walsh; Shaohai Yu

    2003-12-05

    Neural networks were used to calibrate an online ash analyzer at the Usibelli Coal Mine, Healy, Alaska, by relating the Americium and Cesium counts to the ash content. A total of 104 samples were collected from the mine, with 47 being from screened coal, and the rest being from unscreened coal. Each sample corresponded to 20 seconds of coal on the running conveyor belt. Neural network modeling used the quick stop training procedure. Therefore, the samples were split into training, calibration and prediction subsets. Special techniques, using genetic algorithms, were developed to representatively split the sample into the three subsets. Two separate approaches were tried. In one approach, the screened and unscreened coal was modeled separately. In another, a single model was developed for the entire dataset. No advantage was seen from modeling the two subsets separately. The neural network method performed very well on average but not individually, i.e. though each prediction was unreliable, the average of a few predictions was close to the true average. Thus, the method demonstrated that the analyzers were accurate at 2-3 minutes intervals (average of 6-9 samples), but not at 20 seconds (each prediction).

  9. Energy calibration of very inclined air showers

    Energy Technology Data Exchange (ETDEWEB)

    Dembinski, Hans [Institut fuer Experimentelle Kernphysik, Karlsruher Institut fuer Technologie (KIT), Karlsruhe (Germany); Roth, Markus [Institut fuer Kernphysik, Karlsruher Institut fuer Technologie (KIT), Karlsruhe (Germany); Hebbeker, Thomas [III. Physikalisches Institut A, RWTH Aachen (Germany)

    2010-07-01

    The Pierre Auger Observatory detects extensive air showers which are initiated by ultra-high energy cosmic rays. The properties of the cosmic rays are derived indirectly from the air shower observation. The surface detector of the observatory is well suited to detect showers with zenith angles from 0 to 90 . Standard analyses focus on so called vertical showers with inclinations smaller than 60 . Showers with larger zenith angles are called very inclined showers. Both have distinct experimental signatures which require separate event reconstructions. The ground signal of very inclined air showers is muon dominated and the energy reconstruction uses the variable R{sub {mu}} as an estimator for the cosmic ray energy which is proportional to the total number of muons N{sub {mu}} on the ground. The talk will focus on the energy calibration of R{sub {mu}} with events observed simultaneously in the surface and fluorescence detector of the observatory. This calibration procedure also offers the unique opportunity to derive the shower-to-shower fluctuations of R{sub {mu}} which are sensitive to the cosmic ray mass composition.

  10. Calibration of action cameras for photogrammetric purposes.

    Science.gov (United States)

    Balletti, Caterina; Guerra, Francesco; Tsioukas, Vassilios; Vernier, Paolo

    2014-09-18

    The use of action cameras for photogrammetry purposes is not widespread due to the fact that until recently the images provided by the sensors, using either still or video capture mode, were not big enough to perform and provide the appropriate analysis with the necessary photogrammetric accuracy. However, several manufacturers have recently produced and released new lightweight devices which are: (a) easy to handle, (b) capable of performing under extreme conditions and more importantly (c) able to provide both still images and video sequences of high resolution. In order to be able to use the sensor of action cameras we must apply a careful and reliable self-calibration prior to the use of any photogrammetric procedure, a relatively difficult scenario because of the short focal length of the camera and its wide angle lens that is used to obtain the maximum possible resolution of images. Special software, using functions of the OpenCV library, has been created to perform both the calibration and the production of undistorted scenes for each one of the still and video image capturing mode of a novel action camera, the GoPro Hero 3 camera that can provide still images up to 12 Mp and video up 8 Mp resolution.

  11. Calibration of Action Cameras for Photogrammetric Purposes

    Directory of Open Access Journals (Sweden)

    Caterina Balletti

    2014-09-01

    Full Text Available The use of action cameras for photogrammetry purposes is not widespread due to the fact that until recently the images provided by the sensors, using either still or video capture mode, were not big enough to perform and provide the appropriate analysis with the necessary photogrammetric accuracy. However, several manufacturers have recently produced and released new lightweight devices which are: (a easy to handle, (b capable of performing under extreme conditions and more importantly (c able to provide both still images and video sequences of high resolution. In order to be able to use the sensor of action cameras we must apply a careful and reliable self-calibration prior to the use of any photogrammetric procedure, a relatively difficult scenario because of the short focal length of the camera and its wide angle lens that is used to obtain the maximum possible resolution of images. Special software, using functions of the OpenCV library, has been created to perform both the calibration and the production of undistorted scenes for each one of the still and video image capturing mode of a novel action camera, the GoPro Hero 3 camera that can provide still images up to 12 Mp and video up 8 Mp resolution.

  12. 12 CFR 611.510 - Approval procedures.

    Science.gov (United States)

    2010-01-01

    ... 12 Banks and Banking 6 2010-01-01 2010-01-01 false Approval procedures. 611.510 Section 611.510 Banks and Banking FARM CREDIT ADMINISTRATION FARM CREDIT SYSTEM ORGANIZATION Transfer of Authorities § 611.510 Approval procedures. (a) Upon receipt of approval of a resolution by the Farm...

  13. Solar Cell Calibration and Measurement Techniques

    Science.gov (United States)

    Bailey, Sheila; Brinker, Dave; Curtis, Henry; Jenkins, Phillip; Scheiman, Dave

    2004-01-01

    The increasing complexity of space solar cells and the increasing international markets for both cells and arrays has resulted in workshops jointly sponsored by NASDA, ESA and NASA. These workshops are designed to obtain international agreement on standardized values for the AMO spectrum and constant, recommend laboratory measurement practices and establish a set of protocols for international comparison of laboratory measurements. A working draft of an ISO standard, WD15387, "Requirements for Measurement and Calibration Procedures for Space Solar Cells" was discussed with a focus on the scope of the document, a definition of primary standard cell, and required error analysis for all measurement techniques. Working groups addressed the issues of Air Mass Zero (AMO) solar constant and spectrum, laboratory measurement techniques, and te international round robin methodology. A summary is presented of the current state of each area and the formulation of the ISO document.

  14. Calibration of radon-222 detectors using closed circuit radium-226 sources; Calibracao de detectores de radonio-222 atraves do uso de fontes de radio-222 em circuito fechado

    Energy Technology Data Exchange (ETDEWEB)

    Perna, Allan Felipe Nunes; Paschuk, Sergei Anatolyevich; Correa, Janine Nicolosi; Del Claro, Flavia, E-mail: allan_perna@hotmail.com, E-mail: sergei@utfpr.edu.br, E-mail: janine@utfpr.edu.br, E-mail: flavia_delclaro@yahoo.com.br [Universidade Tecnologica Federal do Parana (UTFPR), Curitiba, PR (Brazil)

    2012-07-01

    This paper presents the results of the calibration of the Radon-222 detectors used by the Laboratories specializing in measuring natural radiation from this gas. The research was conducted in collaboration between UTFPR, CDTN/CNEN, UFRN and IRD/CNEN. During the calibration the detectors were exposed in isolated chambers with radioactive calibrated sources. The calibration procedure was supported with four instant radon monitors AlphaGUARD (SAPHYMO Co.) responsible for radon activity measurements in the experimental chamber. The calibration procedure resulted an equation that relates the number of tracks found in solid-state detector CR-39 (Track-Etch detector) with the concentration of radon in the atmosphere. Obtained results are compatible with previously performed calibration at the National Institute of Radiological Sciences (NIRS, Japan) using high activity levels of radon in air. Present results of calibration give the possibility to expand the calibration curve of CR-39 for medium and low activity levels of radon. (author)

  15. Absolute calibration of the colour index and O4 absorption derived from Multi AXis (MAX-)DOAS measurements and their application to a standardised cloud classification algorithm

    Science.gov (United States)

    Wagner, Thomas; Beirle, Steffen; Remmers, Julia; Shaiganfar, Reza; Wang, Yang

    2016-09-01

    A method is developed for the calibration of the colour index (CI) and the O4 absorption derived from differential optical absorption spectroscopy (DOAS) measurements of scattered sunlight. The method is based on the comparison of measurements and radiative transfer simulations for well-defined atmospheric conditions and viewing geometries. Calibrated measurements of the CI and the O4 absorption are important for the detection and classification of clouds from MAX-DOAS observations. Such information is needed for the identification and correction of the cloud influence on Multi AXis (MAX-)DOAS profile inversion results, but might be also be of interest on their own, e.g. for meteorological applications. The calibration algorithm was successfully applied to measurements at two locations: Cabauw in the Netherlands and Wuxi in China. We used CI and O4 observations calibrated by the new method as input for our recently developed cloud classification scheme and also adapted the corresponding threshold values accordingly. For the observations at Cabauw, good agreement is found with the results of the original algorithm. Together with the calibration procedure of the CI and O4 absorption, the cloud classification scheme, which has been tuned to specific locations/conditions so far, can now be applied consistently to MAX-DOAS measurements at different locations. In addition to the new threshold values, further improvements were introduced to the cloud classification algorithm, namely a better description of the SZA (solar zenith angle) dependence of the threshold values and a new set of wavelengths for the determination of the CI. We also indicate specific areas for future research to further improve the cloud classification scheme.

  16. Geometric calibration of the SND detector electromagnetic calorimeter

    Science.gov (United States)

    Korol, A. A.; Melnikova, N. A.

    2017-03-01

    This paper presents the design, implementation and validation of the software alignment procedure used to perform geometric calibration of the electromagnetic calorimeter with respect to the tracking system of the SND detector which is taking data at the VEPP-2000 e+e- collider (BINP, Novosibirsk). This procedure is based on the mathematical model describing the relative calorimeter position. The parameter values are determined by minimizing a χ2 function using the difference between particle directions reconstructed in these two subdetectors for e+e- →e+e- scattering events. The results of the calibration applied to data and MC simulation fit the model well and give an improvement in particle reconstruction. They are used in data reconstruction and MC simulation.

  17. Calibration of the Capintec CRC-712M dose calibrator for (18)F.

    Science.gov (United States)

    Mo, L; Reinhard, M I; Davies, J B; Alexiev, D; Baldock, C

    2006-04-01

    Primary standardisation was performed on a solution of (18)F using the 4pibeta-gamma coincidence counting efficiency-tracing extrapolation method with (60)Co used as a tracer nuclide. The result was used to calibrate the ANSTO secondary standard ionisation chamber which is used to disseminate Australian activity standards for gamma emitters. Using the secondary activity standard for (18)F, the Capintec CRC-712M dose calibrator at the Australian National Medical Cyclotron (NMC) Positron Emission Tomography (PET) Quality Control (QC) Section was calibrated. The dial setting number recommended by the manufacturer for the measurement of the activity of (18)F is 439. In this work, the dial setting numbers for the activity measurement of the solution of (18)F in Wheaton vials were experimentally determined to be 443+/-12, 446+/-12, 459+/-11, 473+/-15 for 0.1, 1, 4.5 and 9ml solution volumes, respectively. The uncertainties given above are expanded uncertainties (k=2) giving an estimated level of confidence of 95%. The activities determined using the manufacturer recommended setting number 439 are 0.8%, 1.4%, 4.0% and 6.5% higher than the standardised activities, respectively. It is recommended that a single dial setting number of 459 determined for 4.5ml is used for 0.1-9ml solution in Wheaton vials in order to simplify the operation procedure. With this setting the expended uncertainty (k=2) in the activity readout from the Capintec dose calibrator would be less than 6.2%.

  18. Payment Procedures for Electronic Government Services

    OpenAIRE

    Pousttchi, Key; Wiedemann, Dietmar Georg

    2005-01-01

    In this paper, we analyse payment procedures for their suitability for electronic government. We provide an overview of the payment procedures currently available on the market, compare the situation in electronic government with the situation in electronic commerce and analyze what we can transfer. Based on functional requirements we propose a scheme that allows public agencies to examine the appropriateness of any given payment procedure for electronic government.

  19. Kali: The framework for fine calibration of LHCb Electromagnetic Calorimeter

    CERN Document Server

    Puig, A; Graciani, R; Belyaev, I

    2011-01-01

    The precise calibration (at a level of below 1 %) of the electromagnetic calorimeter (ECAL) of the LHCb experiment is an essential task for the fulfilment of the LHCb physics program. The final step of this task is performed with two calibration methods using the real data from the experimental setup. It is a very CPU-consuming procedure as both methods require processing of $\\mathcal{O}10^8$ events which must be selected, reconstructed and analyzed. In this document we present the Kali framework developed within the LHCb software framework, which implements these two final calibration methods. It is integrated with Grid middleware and makes use of parallelism tools, such as python parallel processing modules, to provide an efficient way, both time and disk wise, for the final ECAL calibration. The results of the fine calibration with the very first data collected by the LHCb experiment will also be presented. With the use of the Kali framework it took only two days of processing and allowed to achieve a cali...

  20. IMU-Based Online Kinematic Calibration of Robot Manipulator

    Directory of Open Access Journals (Sweden)

    Guanglong Du

    2013-01-01

    Full Text Available Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA and Kalman Filter (KF to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  1. IMU-based online kinematic calibration of robot manipulator.

    Science.gov (United States)

    Du, Guanglong; Zhang, Ping

    2013-01-01

    Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU) is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA) and Kalman Filter (KF) to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF) is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  2. Application of the SKYRAD Improved Langley plot method for the in situ calibration of CIMEL Sun-sky photometers.

    Science.gov (United States)

    Campanelli, Monica; Estellés, Víctor; Tomasi, Claudio; Nakajima, Teruyuki; Malvestuto, Vincenzo; Martínez-Lozano, José Antonio

    2007-05-10

    The in situ procedure for determining the solar calibration constants, originally developed for the PREDE Sun-sky radiometers and based on a modified version of the Langley plot, was applied to a CIMEL instrument located in Valencia, Spain, not integrated into AERONET. Taking into account the different mechanical and electronic characteristics of the two radiometers, the method was adapted to the characteristics of the CIMEL instrument. The iterative procedure for the determination of the solar calibration constants was applied to a 3-year data set. The results were compared with the two sets of experimental calibration constants determined during this period using the standard Langley plot method. The agreement was found to be consistent with the experimental errors, and the method can definitely also be used to determine the solar calibration constant for the CIMEL instrument, improving its calibration. The method can be used provided the radiometer is previously calibrated for diffuse radiance using a standard lamp.

  3. User-calibration of Fowler Ultra-Cal Mark III Digital caliper

    Energy Technology Data Exchange (ETDEWEB)

    Estill, J.

    1996-09-19

    The purpose of this technical implementing procedure (TIP) is to describe the procedure that will be employed for user-calibration of a digital caliper used in the determination of specimen dimensions. A caliper is used for some of the activities of the Scientific Investigation Plan (SIP) Metal Barrier Selection and Testing (SIP-CM-01, WBS {number_sign} 1.2.2.5.1). In particular, it will be used for Activity E-20-50, Long-Term Corrosion Studies. This procedure describes the methodology for user calibration of a Fowler Ultra-Cal Mark III digital caliper. National Institutes of Standards and Technology (NIST) traceable gauge blocks are employed in the calibration procedure.

  4. Liquid Scintillation Counting - Packard Triple-Label Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Torretto, P. A. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2017-03-23

    The Radiological Measurements Laboratory (RML) maintains and operates nine Packard Liquid Scintillation Counters (LSCs). These counters were obtained through various sources and were generally purchased as 2500, 2700 or 3100 series counters. In 2004/2005 the software and firmware on the counters were upgraded. The counters are now designated as 3100 series counters running the Quantasmart software package. Thus, a single procedure can be used to calibrate and operate the Packard LSCs.

  5. The role of entanglement in calibrating optical quantum gyroscopes

    OpenAIRE

    Kok, Pieter; Dunningham, Jacob; Ralph, Jason F.

    2015-01-01

    We consider the calibration of an optical quantum gyroscope by modeling two Sagnac interferometers, mounted approximately at right angles to each other. Reliable operation requires that we know the angle between the interferometers with high precision, and we show that a procedure akin to multi-position testing in inertial navigation systems can be generalized to the case of quantum interferometry. We find that while entanglement is a key resource within an individual Sagnac interferometer, i...

  6. A self-sufficient method for calibration of Varian electronic portal imaging device

    Science.gov (United States)

    Sun, Baozhou; Yaddanapudi, Sridhar; Goddu, Sreekrishna M.; Mutic, Sasa

    2015-01-01

    Electronic portal imaging device (EPID) is currently used for dosimetric verification of IMRT fields and linac quality assurance (QA). It is critical to understand the dosimetric response and perform an accurate and robust calibration of EPID. We present the implementation of an efficient method for the calibration and the validation of a Varian EPID, which relies only on data collected with that specific device. The calibration method is based on images obtained with five shifts of EPID panel. With this method, the relative gain (sensitivity) of each element of a detector matrix is calculated and applied on top of the calibration determined with the flood-field procedure. The calibration procedure was verified using a physical wedge inserted in the beam line and the corrected profile shows consistent results with the measurements using a calibrated 2D array. This method does not rely on the beam profile used in the flood-field calibration process, which allows EPID calibration in 10 minutes with no additional equipment compared to at least 2 hours to obtain beam profile and scanning beam equipment requirement with the conventional method.

  7. Reduced Ambiguity Calibration for LOFAR

    CERN Document Server

    Yatawatta, Sarod

    2012-01-01

    Interferometric calibration always yields non unique solutions. It is therefore essential to remove these ambiguities before the solutions could be used in any further modeling of the sky, the instrument or propagation effects such as the ionosphere. We present a method for LOFAR calibration which does not yield a unitary ambiguity, especially under ionospheric distortions. We also present exact ambiguities we get in our solutions, in closed form. Casting this as an optimization problem, we also present conditions for this approach to work. The proposed method enables us to use the solutions obtained via calibration for further modeling of instrumental and propagation effects. We provide extensive simulation results on the performance of our method. Moreover, we also give cases where due to degeneracy, this method fails to perform as expected and in such cases, we suggest exploiting diversity in time, space and frequency.

  8. Astrid-2 EMMA Magnetic Calibration

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Risbo, Torben

    1998-01-01

    The Swedish micro-satellite Astrid-2 contains a tri-axial fluxgate magnetometer with the sensor co-located with a Technical University of Denmark (DTU) star camera for absolute attitude, and extended about 0.9 m on a hinged boom. The magnetometer is part of the RIT EMMA electric and magnetic fields...... Survey of Sweden near Stockholm on the night of May 15.-16., 1997. The magnetic calibration and the intercalibration between the star camera and the magnetic sensor was performed by measuring the Earth's magnetic field and simultaneously observing the star sky with the camera. The rotation matrix between...... fit calibration parameters. Owing to time shortage, we did not evaluate the temperature coefficients of the flight sensor calibration parameters. However, this was done for an identical flight spare magnetometer sensor at the magnetic coil facility belonging to the Technical University of Braunschweig...

  9. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    of non-uniformity of the illumination of the image plane. Only the image deforming aberrations and the non-uniformity of illumination are included in the calibration models. The topics of the pinhole camera model and the extension to the Direct Linear Transform (DLT) are described. It is shown how......The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... the DLT can be extended with non-linear models of the common lens aberrations/errors some of them caused by manufacturing defects like decentering and thin prism distortion. The relation between a warping and the non-linear defects are shown. The issue of making a good resampling of an image by using...

  10. Balanced calibration of resonant shunt circuits for piezoelectric vibration control

    DEFF Research Database (Denmark)

    Høgsberg, Jan; Krenk, Steen

    2012-01-01

    Shunting of piezoelectric transducers and suitable electric circuits constitutes an effective passive approach to resonant vibration damping of structures. Most common design concepts for resonant resistor-inductor (RL) shunt circuits rely on either maximization of the attainable modal damping...... series and parallel RL circuits. The procedure relies on equal modal damping and sufficient separation of the complex poles to avoid constructive interference of the two modes. By comparison with existing design procedures, it is demonstrated that the present calibration leads to a balanced compromise...

  11. Calibration and Measurement Uncertainty Estimation of Radiometric Data: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, A.; Sengupta, M.; Reda, I.; Andreas, A.; Konings, J.

    2014-11-01

    Evaluating the performance of photovoltaic cells, modules, and arrays that form large solar deployments relies on accurate measurements of the available solar resource. Therefore, determining the accuracy of these solar radiation measurements provides a better understanding of investment risks. This paper provides guidelines and recommended procedures for estimating the uncertainty in calibrations and measurements by radiometers using methods that follow the International Bureau of Weights and Measures Guide to the Expression of Uncertainty (GUM). Standardized analysis based on these procedures ensures that the uncertainty quoted is well documented.

  12. Tank calibration; Arqueacao de tanques

    Energy Technology Data Exchange (ETDEWEB)

    Chan, Ana [PETROBRAS S.A., Rio de Janeiro, RJ (Brazil)

    2003-07-01

    This work relates the analysis of the norms ISO (International Organization for Standardization) for calibration of vertical cylindrical tanks used in fiscal measurement, established on Joint Regulation no 1 of June 19, 2000 between the ANP (National Agency of Petroleum) and the INMETRO (National Institute of Metrology, Normalization and Industrial Quality). In this work a comparison between norms ISO and norms published by the API (American Petroleum Institute) and the IP (Institute of Petroleum) up to 2001 was made. It was concluded that norms ISO are wider than norms API, IP, and INMETRO methods in the calibration of vertical cylindrical tanks. (author)

  13. Performance standard for dose Calibrator

    CERN Document Server

    Darmawati, S

    2002-01-01

    Dose calibrator is an instrument used in hospitals to determine the activity of radionuclide for nuclear medicine purposes. International Electrotechnical Commission (IEC) has published IEC 1303:1994 standard that can be used as guidance to test the performance of the instrument. This paper briefly describes content of the document,as well as explains the assessment that had been carried out to test the instrument accuracy in Indonesia through intercomparison measurement.Its is suggested that hospitals acquire a medical physicist to perform the test for its dose calibrator. The need for performance standard in the form of Indonesia Standard is also touched.

  14. Analytical Chemistry Laboratory (ACL) procedure compendium. Volume 7, Safety operation procedure for hot cell

    Energy Technology Data Exchange (ETDEWEB)

    1993-08-01

    This volume contains the interim change notice for the safety operation procedure for hot cell. It covers the master-slave manipulators, dry waste removal, cell transfers, hoists, cask handling, liquid waste system, and physical characterization of fluids.

  15. Shortwave Radiometer Calibration Methods Comparison and Resulting Solar Irradiance Measurement Differences: A User Perspective

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-11-21

    Banks financing solar energy projects require assurance that these systems will produce the energy predicted. Furthermore, utility planners and grid system operators need to understand the impact of the variable solar resource on solar energy conversion system performance. Accurate solar radiation data sets reduce the expense associated with mitigating performance risk and assist in understanding the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of different calibration methods provided by radiometric calibration service providers, such as NREL and manufacturers of radiometers, on the resulting calibration responsivity. Some of these radiometers are calibrated indoors and some outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The BORCAL method provides the outdoor calibration responsivity of pyranometers and pyrheliometers at 45 degree solar zenith angle, and as a function of solar zenith angle determined by clear-sky comparisons with reference irradiance. The BORCAL method also employs a thermal offset correction to the calibration responsivity of single-black thermopile detectors used in pyranometers. Indoor calibrations of radiometers by their manufacturers are performed using a stable artificial light source in a side-by-side comparison between the test radiometer under calibration and a reference radiometer of the same type. In both methods, the reference radiometer calibrations are traceable to the World Radiometric Reference (WRR). These

  16. Principles of tendon transfers.

    Science.gov (United States)

    Coulet, B

    2016-04-01

    Tendon transfers are carried out to restore functional deficits by rerouting the remaining intact muscles. Transfers are highly attractive in the context of hand surgery because of the possibility of restoring the patient's ability to grip. In palsy cases, tendon transfers are only used when a neurological procedure is contraindicated or has failed. The strategy used to restore function follows a common set of principles, no matter the nature of the deficit. The first step is to clearly distinguish between deficient muscles and muscles that could be transferred. Next, the type of palsy will dictate the scope of the program and the complexity of the gripping movements that can be restored. Based on this reasoning, a surgical strategy that matches the means (transferable muscles) with the objectives (functions to restore) will be established and clearly explained to the patient. Every paralyzed hand can be described using three parameters. 1) Deficient segments: wrist, thumb and long fingers; 2) mechanical performance of muscles groups being revived: high energy-wrist extension and finger flexion that require strong transfers with long excursion; low energy-wrist flexion and finger extension movements that are less demanding mechanically, because they can be accomplished through gravity alone in some cases; 3) condition of the two primary motors in the hand: extrinsics (flexors and extensors) and intrinsics (facilitator). No matter the type of palsy, the transfer surgery follows the same technical principles: exposure, release, fixation, tensioning and rehabilitation. By performing an in-depth analysis of each case and by following strict technical principles, tendon transfer surgery leads to reproducible results; this allows the surgeon to establish clear objectives for the patient preoperatively.

  17. Third-party brachytherapy source calibrations and physicist responsibilities: report of the AAPM Low Energy Brachytherapy Source Calibration Working Group.

    Science.gov (United States)

    Butler, Wayne M; Bice, William S; DeWerd, Larry A; Hevezi, James M; Huq, M Saiful; Ibbott, Geoffrey S; Palta, Jatinder R; Rivard, Mark J; Seuntjens, Jan P; Thomadsen, Bruce R

    2008-09-01

    The AAPM Low Energy Brachytherapy Source Calibration Working Group was formed to investigate and recommend quality control and quality assurance procedures for brachytherapy sources prior to clinical use. Compiling and clarifying recommendations established by previous AAPM Task Groups 40, 56, and 64 were among the working group's charges, which also included the role of third-party handlers to perform loading and assay of sources. This document presents the findings of the working group on the responsibilities of the institutional medical physicist and a clarification of the existing AAPM recommendations in the assay of brachytherapy sources. Responsibility for the performance and attestation of source assays rests with the institutional medical physicist, who must use calibration equipment appropriate for each source type used at the institution. Such equipment and calibration procedures shall ensure secondary traceability to a national standard. For each multi-source implant, 10% of the sources or ten sources, whichever is greater, are to be assayed. Procedures for presterilized source packaging are outlined. The mean source strength of the assayed sources must agree with the manufacturer's stated strength to within 3%, or action must be taken to resolve the difference. Third party assays do not absolve the institutional physicist from the responsibility to perform the institutional measurement and attest to the strength of the implanted sources. The AAPM leaves it to the discretion of the institutional medical physicist whether the manufacturer's or institutional physicist's measured value should be used in performing dosimetry calculations.

  18. 40 CFR 86.120-94 - Gas meter or flow instrumentation calibration; particulate, methanol and formaldehyde measurement.

    Science.gov (United States)

    2010-07-01

    ... calibration; particulate, methanol and formaldehyde measurement. 86.120-94 Section 86.120-94 Protection of... Procedures § 86.120-94 Gas meter or flow instrumentation calibration; particulate, methanol and formaldehyde measurement. (a) Sampling for particulate, methanol and formaldehyde emissions requires the use of gas...

  19. Operation and calibration of the Silicon Drift Detectors of the ALICE experiment during the 2008 cosmic ray data taking period

    CERN Document Server

    Alessandro, B; Bala, R; Batigne, G; Beolè, S; Biolcati, E; Bock Garcia, N; Bruna, E; Cerello, P; Coli, S; Corrales Morales, Y; Costa, F; Crescio, E; De Remigis, P; Di Liberto, S; Falchieri, D; Feofilov, G; Ferrarese, W; Gandolfi, E; Garcia, C; Gaudichet, L; Giraudo, G; Giubellino, P; Humanic, T J; Igolkin, S; Idzik, M; Kiprich, S K; Kisiel, A; Kolozhvari, A; Kotov, I; Kral, J; Kushpil, S; Kushpil, V; Lea, R; Lisa, M A; Martinez, M I; Marzari Chiesa, A; Masera, M; Masetti, M; Mazza, G; Mazzoni, M A; Meddi, F; Montano Zetina, L M; Monteno, M; Nilsen, B S; Nouais, D; Padilla Cabal, F; Petrácek, V; Poghosyan, M G; Prino, F; Ramello, L; Rashevsky, A; Riccati, L; Rivetti, A S; Senyukov, S; Siciliano, M; Sitta, M; Subieta Vasquez, M A; Sumbera, M L; Toscano, L; Tosello, F; Truesdale, D; Urciuoli, G M; Vacchi, A; Vallero, S; Werbrouck, A; Zampa, G; Zinovjev, G

    2010-01-01

    The calibration and performance of the Silicon Drift Detector of the ALICE experiment during the 2008 cosmic ray run will be presented. In particular the procedures to monitor the running parameters (baselines, noise, drift speed) are detailed. Other relevant parameters (SOP delay, time-zero, charge calibration) were also determined.

  20. Balanced calibration of resonant piezoelectric RL shunts with quasi-static background flexibility correction

    DEFF Research Database (Denmark)

    Høgsberg, Jan Becker; Krenk, Steen

    2015-01-01

    Resonant RL shunt circuits constitute a robust approach to piezoelectric damping, where the performance with respect to damping of flexible structures requires a precise calibration of the corresponding circuit components. The balanced calibration procedure of the present paper is based on equal ...

  1. A likelihood method to cross-calibrate air-shower detectors

    CERN Document Server

    Dembinski, H P; Mariş, I C; Roth, M; Veberič, D

    2016-01-01

    We present a detailed statistical treatment of the energy calibration of hybrid air-shower detectors, which combine a surface detector array and a fluorescence detector, to obtain an unbiased estimate of the calibration curve. The special features of calibration data from air showers prevent unbiased results, if a standard least-squares fit is applied to the problem. We develop a general maximum-likelihood approach, based on the detailed statistical model, to solve the problem. Our approach was developed for the Pierre Auger Observatory, but the applied principles are general and can be transferred to other air-shower experiments, even to the cross-calibration of other observables. Since our general likelihood function is expensive to compute, we derive two approximations with significantly smaller computational cost. In the recent years both have been used to calibrate data of the Pierre Auger Observatory. We demonstrate that these approximations introduce negligible bias when they are applied to simulated t...

  2. Transfer Pricing

    DEFF Research Database (Denmark)

    Rohde, Carsten; Rossing, Christian Plesner

    trade internally as the units have to decide what prices should be paid for such inter-unit transfers. One important challenge is to uncover the consequences that different transfer prices have on the willingness in the organizational units to coordinate activities and trade internally. At the same time...

  3. Drift-insensitive distributed calibration of probe microscope scanner in nanometer range: Virtual mode

    Science.gov (United States)

    Lapshin, Rostislav V.

    2016-08-01

    A method of distributed calibration of a probe microscope scanner is suggested. The main idea consists in a search for a net of local calibration coefficients (LCCs) in the process of automatic measurement of a standard surface, whereby each point of the movement space of the scanner can be characterized by a unique set of scale factors. Feature-oriented scanning (FOS) methodology is used as a basis for implementation of the distributed calibration permitting to exclude in situ the negative influence of thermal drift, creep and hysteresis on the obtained results. Possessing the calibration database enables correcting in one procedure all the spatial systematic distortions caused by nonlinearity, nonorthogonality and spurious crosstalk couplings of the microscope scanner piezomanipulators. To provide high precision of spatial measurements in nanometer range, the calibration is carried out using natural standards - constants of crystal lattice. One of the useful modes of the developed calibration method is a virtual mode. In the virtual mode, instead of measurement of a real surface of the standard, the calibration program makes a surface image "measurement" of the standard, which was obtained earlier using conventional raster scanning. The application of the virtual mode permits simulation of the calibration process and detail analysis of raster distortions occurring in both conventional and counter surface scanning. Moreover, the mode allows to estimate the thermal drift and the creep velocities acting while surface scanning. Virtual calibration makes possible automatic characterization of a surface by the method of scanning probe microscopy (SPM).

  4. Scalar Calibration of Vector Magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz;

    2000-01-01

    The calibration parameters of a vector magnetometer are estimated only by the use of a scalar reference magnetometer. The method presented in this paper differs from those previously reported in its linearized parametrization. This allows the determination of three offsets or signals in the absence...

  5. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Villanueva, Héctor

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this repor...

  6. CALIBRATION OF THE INFRARED OPTOMETER

    Science.gov (United States)

    An infrared optometer for measuring the absolute status of accommodation is subject to a constant error not associated with chromatic aberration or...on optometer accuracy as long as the pupil does not vignette the optometer beam. A modification is described for calibrating the infrared optometer ...for an individual subject without using trial lenses or a subjective optometer . (Author)

  7. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  8. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Vesth, Allan; Kock, Carsten Weber

    The report describes power curve measurements carried out on a given wind turbine. The measurements are carried out in accordance to Ref. [1]. A site calibration has been carried out; see Ref. [2], and the measured flow correction factors for different wind directions are used in the present...

  9. Using standard calibrated geometries to characterize a coaxial high purity germanium gamma detector for Monte Carlo simulations

    NARCIS (Netherlands)

    van der Graaf, E. R.; Dendooven, P.; Brandenburg, S.

    2014-01-01

    A detector model optimization procedure based on matching Monte Carlo simulations with measurements for two experimentally calibrated sample geometries which are frequently used in radioactivity measurement laboratories results in relative agreement within 5% between simulated and measured efficienc

  10. Towards a global network of gamma-ray detector calibration facilities

    Science.gov (United States)

    Tijs, Marco; Koomans, Ronald; Limburg, Han

    2016-09-01

    Gamma-ray logging tools are applied worldwide. At various locations, calibration facilities are used to calibrate these gamma-ray logging systems. Several attempts have been made to cross-correlate well known calibration pits, but this cross-correlation does not include calibration facilities in Europe or private company calibration facilities. Our aim is to set-up a framework that gives the possibility to interlink all calibration facilities worldwide by using `tools of opportunity' - tools that have been calibrated in different calibration facilities, whether this usage was on a coordinated basis or by coincidence. To compare the measurement of different tools, it is important to understand the behaviour of the tools in the different calibration pits. Borehole properties, such as diameter, fluid, casing and probe diameter strongly influence the outcome of gamma-ray borehole logging. Logs need to be properly calibrated and compensated for these borehole properties in order to obtain in-situ grades or to do cross-hole correlation. Some tool providers provide tool-specific correction curves for this purpose. Others rely on reference measurements against sources of known radionuclide concentration and geometry. In this article, we present an attempt to set-up a framework for transferring `local' calibrations to be applied `globally'. This framework includes corrections for any geometry and detector size to give absolute concentrations of radionuclides from borehole measurements. This model is used to compare measurements in the calibration pits of Grand Junction, located in the USA; Adelaide (previously known as AMDEL), located in Adelaide Australia; and Stonehenge, located at Medusa Explorations BV in the Netherlands.

  11. Designing Flight Deck Procedures

    Science.gov (United States)

    Degani, Asaf; Wiener, Earl

    2005-01-01

    Three reports address the design of flight-deck procedures and various aspects of human interaction with cockpit systems that have direct impact on flight safety. One report, On the Typography of Flight- Deck Documentation, discusses basic research about typography and the kind of information needed by designers of flight deck documentation. Flight crews reading poorly designed documentation may easily overlook a crucial item on the checklist. The report surveys and summarizes the available literature regarding the design and typographical aspects of printed material. It focuses on typographical factors such as proper typefaces, character height, use of lower- and upper-case characters, line length, and spacing. Graphical aspects such as layout, color coding, fonts, and character contrast are discussed; and several cockpit conditions such as lighting levels and glare are addressed, as well as usage factors such as angular alignment, paper quality, and colors. Most of the insights and recommendations discussed in this report are transferable to paperless cockpit systems of the future and computer-based procedure displays (e.g., "electronic flight bag") in aerospace systems and similar systems that are used in other industries such as medical, nuclear systems, maritime operations, and military systems.

  12. Geo energy research and development: technology transfer

    Energy Technology Data Exchange (ETDEWEB)

    Traeger, R.K.

    1982-03-01

    Sandia Geo Energy Programs related to geothermal, coal, oil and gas, and synfuel resources have provided a useful mechanism for transferring laboratory technologies to private industry. Significant transfer of hardware, computer programs, diagnostics and instrumentation, advanced materials, and in situ process understanding has occurred through US/DOE supported programs in the past five years. The text briefly reviews the technology transfer procedures and summarizes 32 items that have been transferred and another 20 technologies that are now being considered for possible transfer to industry. A major factor in successful transfer has been personal interactions between Sandia engineers and the technical staff from private industry during all aspects of the technology development.

  13. An integrating sphere radiometer as a solution for high power calibrations in fibre optics

    Science.gov (United States)

    Carrasco-Sanz, Ana; Rodríguez-Barrios, Félix; Corredera, Pedro; Martín-López, Sonia; González-Herráez, Miguel; Hernanz, María Luisa

    2006-04-01

    This paper describes the design, characterization and calibration of a high power transfer standard for optical power measurements in optical fibres based on an integrating sphere radiometer. This radiometer, based on two detectors (Si and InGaAs), can measure powers between 100 nW and 10 W within the wavelength range of (400-1700) nm. The radiometer has been calibrated over the total spectral range of use against an electrically calibrated pyroelectric radiometer and different fibre laser diodes and ion lasers. The total uncertainty obtained is lower than ±1.5% for these wavelengths and power ranges (excluding the water absorption region).

  14. LHCb : Novel real-time alignment and calibration of the LHCb Detector in Run2

    CERN Multimedia

    Tobin, Mark

    2015-01-01

    LHCb has introduced a novel real-time detector alignment and calibration strategy for LHC Run 2. Data collected at the start of the fill will be processed in a few minutes and used to update the alignment, while the calibration constants will be evaluated for each run. This procedure will improve the quality of the online alignment. For example, the vertex locator is retracted and reinserted for stable beam collisions in each fill to be centred on the primary vertex position in the transverse plane. Consequently its position changes on a fill-by-fill basis. Critically, this new realtime alignment and calibration procedure allows identical constants to be used in the online and offline reconstruction, thus improving the correlation between triggered and offline selected events. This offers the opportunity to optimise the event selection in the trigger by applying stronger constraints. The online calibration facilitates the use of hadronic particle identification using the RICH detectors at the trigger level. T...

  15. User's manual for MODCAL: Bounding surface soil plasticity model calibration and prediction code, volume 2

    Science.gov (United States)

    Dennatale, J. S.; Herrmann, L. R.; Defalias, Y. F.

    1983-02-01

    In order to reduce the complexity of the model calibration process, a computer-aided automated procedure has been developed and tested. The computer code employs a Quasi-Newton optimization strategy to locate that set of parameter values which minimizes the discrepancy between the model predictions and the experimental observations included in the calibration data base. Through application to a number of real soils, the automated procedure has been found to be an efficient, reliable and economical means of accomplishing model calibration. Although the code was developed specifically for use with the Bounding Surface plasticity model, it can readily be adapted to other constitutive formulations. Since the code greatly reduces the dependence of calibration success on user expertise, it significantly increases the accessibility and usefulness of sophisticated material models to the general engineering community.

  16. Radiation calibration for LWIR Hyperspectral Imager Spectrometer

    Science.gov (United States)

    Yang, Zhixiong; Yu, Chunchao; Zheng, Wei-jian; Lei, Zhenggang; Yan, Min; Yuan, Xiaochun; Zhang, Peizhong

    2014-11-01

    The radiometric calibration of LWIR Hyperspectral imager Spectrometer is presented. The lab has been developed to LWIR Interferometric Hyperspectral imager Spectrometer Prototype(CHIPED-I) to study Lab Radiation Calibration, Two-point linear calibration is carried out for the spectrometer by using blackbody respectively. Firstly, calibration measured relative intensity is converted to the absolute radiation lightness of the object. Then, radiation lightness of the object is is converted the brightness temperature spectrum by the method of brightness temperature. The result indicated †that this method of Radiation Calibration calibration was very good.

  17. Radio Interferometric Calibration Using a Riemannian Manifold

    CERN Document Server

    Yatawatta, Sarod

    2013-01-01

    In order to cope with the increased data volumes generated by modern radio interferometers such as LOFAR (Low Frequency Array) or SKA (Square Kilometre Array), fast and efficient calibration algorithms are essential. Traditional radio interferometric calibration is performed using nonlinear optimization techniques such as the Levenberg-Marquardt algorithm in Euclidean space. In this paper, we reformulate radio interferometric calibration as a nonlinear optimization problem on a Riemannian manifold. The reformulated calibration problem is solved using the Riemannian trust-region method. We show that calibration on a Riemannian manifold has faster convergence with reduced computational cost compared to conventional calibration in Euclidean space.

  18. Improved GPS-based time link calibration involving ROA and PTB.

    Science.gov (United States)

    Esteban, Héctor; Palacio, Juan; Galindo, Francisco Javier; Feldmann, Thorsten; Bauch, Andreas; Piester, Dirk

    2010-03-01

    The calibration of time transfer links is mandatory in the context of international collaboration for the realization of International Atomic Time. In this paper, we present the results of the calibration of the GPS time transfer link between the Real Instituto y Observatorio de la Armada (ROA) and the Physikalisch-Technische Bundesanstalt (PTB) by means of a traveling geodetic-type GPS receiver and an evaluation of the achieved type A and B uncertainty. The time transfer results were achieved by using CA, P3, and also carrier phase PPP comparison techniques. We finally use these results to re-calibrate the two-way satellite time and frequency transfer (TWSTFT) link between ROA and PTB, using one month of data. We show that a TWSTFT link can be calibrated by means of GPS time comparisons with an uncertainty below 2 ns, and that potentially even sub-nanosecond uncertainty can be achieved. This is a novel and cost-effective approach compared with the more common calibration using a traveling TWSTFT station.

  19. Absolute calibration of an ultraviolet spectrometer using a stabilized laser and a cryogenic cavity radiometer

    Science.gov (United States)

    Jauniskis, L.; Foukal, P.; Kochling, H.

    1992-01-01

    We carry out the calibration of an ultraviolet spectrometer by using a cryogenic electrical-substitution radiometer and intensity-stabilized laser sources. A comparison of the error budgets for the laser-based calibration described here and for a calibration using a type-FEL tungsten spectral-irradiance standard indicates that this technique could provide an improvement of a factor of about three in the uncertainty of the spectrometer calibration, resulting in an absolute accuracy (standard deviation of three) of about 1 percent at 257 nm. The technique described here might significantly improve the accuracy of calibrations on NASA ozone-monitoring and solar ultraviolet-monitoring spectrophotometers when used to complement present procedures that employ lamps and the SURF II synchrotron ultraviolet radiation facility at the National Institute of Standards and Technology.

  20. CalVin 3 — A New Release of the ESO Calibrator Selection Tool for the VLT Interferometer

    Science.gov (United States)

    Wittkowski, M.; Ballester, P.; Bonneau, D.; Chelli, A.; Chesneau, O.; Cruzalèbes, P.; Duvert, G.; Hummel, C.; Lafrasse, S.; Mella, G.; Melnick, J.; Mérand, A.; Mourard, D.; Percheron, I.; Sacuto, S.; Shabun, K.; Stefl, S.; Vinther, J.

    2011-09-01

    Interferometric observations require frequent measurements of calibration stars of known diameter to estimate the instrumental transfer function. ESO offers the preparation tool CalVin to select suitable calibrators from an underlying list of calibrators. The latest version 3, first released in January 2011, offers major improvements in the number of available calibrators, the functionality of the search tool, as well as in terms of performance and ease of use. It has been developed in a collaboration between ESO and the French Jean-Marie Mariotti Center (JMMC).