WorldWideScience

Sample records for calibration calibracion del

  1. IIE`s wind tunnel calibration; Calibracion del tunel de viento del IIE

    Energy Technology Data Exchange (ETDEWEB)

    Pena Garcia, Raymundo [Instituto de Investigaciones Electricas, Cuernavaca (Mexico)

    1985-12-31

    The calibration of a wind tunnel is performed in such a way as to warrant a very low turbulence grade. When there is recently built tunnel, as is the case of the IIE`s tunnel, the turbulence in its testing chambers is large; for this reason it is necessary to integrate in it aerodynamic devices and elements capable of reducing it. At the end of the calibration studies can be performed in models with controlled scale. From these and from the results obtained it will be decided if the designed prototypes are built or modified. [Espanol] La calibracion de un tunel de viento se realiza de tal forma que garantiza un grado de turbulencia muy bajo. Cuando se tiene un tunel recien construido, como es el caso del tunel de viento del IIE, la turbulencia en sus camaras de prueba es grande; por lo que es necesario integrarle dispositivos y elementos aerodinamicos que sean capaces de reducirla. Al terminar la calibracion pueden realizarse estudios en modelos con escala controlada. De estos y de los resultados que se obtengan se decidira si se construyen o se modifican los prototipos disenados.

  2. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR; Automatizacion de la calibracion de dosimetros de radioterapia en el laboratorio secundario de calibracion dosimetrica del CPHR

    Energy Technology Data Exchange (ETDEWEB)

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez, E-mail: c19btm@frcuba.co.cu [Centro de Proteccion e Higiene de las Radiaciones (CPHR), La Habana (Cuba)

    2013-11-01

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba.

  3. Calibration of the indicators of the product KERMa-AREA in radiodiagnostic equipment; Calibracion de los indicadores del producto KERMA-AREA en los equipos de radiodiagnostico

    Energy Technology Data Exchange (ETDEWEB)

    Ginjaume, M.; Jarvinen, H.; Turak, O.

    2013-07-01

    Currently the IAEA proposes two procedures for the calibration of the determination of the PKA systems integrated in the x-ray equipment. The traditional method is to measure air kerma at a certain distance from the focus through an ionization chamber reference and then the area of the field of radiation in that position. The alternative method is based on measuring directly the product kerma-area using equipment calibrated in this magnitude. In both cases the calibration procedure takes into account the position of the tube on the stretcher of the patient. The objective of this study was to compare the practical application of both calibration procedures as well as assess the accuracy of indication of the PKA in the consoles of the radiology equipment. (Author)

  4. Calibration and start-up of the system of radiotherapy extract image-guided; Calibracion y puesta en marcha del sistema de radioterpia guiada por la imagen exactrac

    Energy Technology Data Exchange (ETDEWEB)

    Clemente Gutierrez, I.; Perez Vara, C.; Prieto Villacorta, M.

    2013-07-01

    The accuracy in the administration of external radiotherapy treatments may be increased through the use of guidance systems by image. Constitute a particular case x-rays teams independent of the treatment unit. Among them is the ExacTrac (Brainlab) system. The objective of this work is to briefly introduce the procedure followed in the calibration and implementation of such a system. (Author)

  5. Dosimetry through the Secondary Laboratory of Dosimetric Calibration of Mexico; Dosimetria a traves del Laboratorio Secundario de Calibracion Dosimetrica de Mexico

    Energy Technology Data Exchange (ETDEWEB)

    Tovar M, V.M.; Alvarez R, J.T.; Medina O, V.P.; Vergara M, F.; Anaya M, R.; Cejudo A, J.; Salinas L, B. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2004-07-01

    In the beginnings of the sixty years an urgent necessity is presented mainly in the developing countries, of improving in important form the accuracy in the dosimetry of external faces in therapy of radiations (radiotherapy centers), mainly in the calibration of ''clinical dosemeters''. In 1976 the International Atomic Energy Agency, (IAEA), and the World Health Organization, (WHO), they carried out a mutual agreement with regard to the establishment and operation of a net of Secondary Patron Laboratories of Dosimetry, (LSCD). The necessity to establish measure patterns in the field of the dosimetry of the ionizing radiations, is necessary, to have an accuracy but high in the dosimetry of the radiation beams in therapy which is highly dependent of the dose given to the tumor of those patient with cancer. Similar levels of accuracy are required in protection measures to the radiation with an acceptable smaller accuracy, however, when the personal dosemeters are used to determine the doses received by the individuals under work conditions, such mensurations in therapy of radiations and radiological protection will have traceability through a chain of comparisons to primary or national patterns. The traceability is necessary to assure the accuracy and acceptability of the dosimetric measures, as well as, the legal and economic implications. The traceability is also necessary in the dosimetry of high dose like in the sterilization of different products. The main function of the LSCD is to provide a service in metrology of ionizing radiations, maintaining the secondary or national patterns, which have a traceability to the International System of measures, which is based for if same in the comparison of patterns in the Primary Laboratories of Dosimetry (LPD) under the auspice of the International Office of Weights and Measure (BIPM). The secondary and national patterns in the LSCD constitute in Mexico, the national patterns of the magnitudes in the

  6. Characterization of the neutron field of the {sup 241}AmBe in a calibration room; Caracterizacion del campo de neutrones del {sup 241} AmBe en una sala para calibracion

    Energy Technology Data Exchange (ETDEWEB)

    Vega C, H.R. [UAZ, A.P. 336, 98000 Zacatecas (Mexico); Gallego, E.; Lorente, A. [Universidad Politecnica de Madrid, C. Jose Gutierrez Abascal 2, 28006 Madrid (Spain)] e-mail: rvega@cantera.reduaz.mx

    2003-07-01

    The field of neutrons produced by an isotopic source of neutrons of {sup 241} Am Be had been characterized. The characterization was carried out modeling those relevant details of the calibration room and simulating the neutron transport at different distances of the source. The calculated spectra were used to determine the equivalent environmental dose rate. A series of experiments were carried out with the Bonner sphere spectrometric system to measure the spectra in the same points where the calculations were carried out and with these spectra the rates of environmental dose were calculated. By means of a one sphere dosemeter type Berthold the rates of environmental dose were measured. To the one to compare the calculated spectra and measured its were found small differences in the group of the thermal neutrons due to the elementary composition used during the simulation. When comparing the derived rates starting from the calculated spectra with those measured it was found a maxim difference smaller to 13%. (Author)

  7. Parameters of calibration of the measurement system of {sup 222} Rn based in LR-115; Parametros de calibracion del sistema de medicion de {sup 222} Rn basado en LR-115

    Energy Technology Data Exchange (ETDEWEB)

    Garcia, M.L.; Mireles, F.; Quirino, L.; Davila, I.; Lugo, F.; Pinedo, J.L. [CREN-UAZ, 98068 Zacatecas (Mexico); Chavez, A. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)]. e-mail: mluisagb@hotmail.com

    2003-07-01

    Since the SSNTD technique (Solid State Nuclear Track Detection) it was discovered it has been used as passive method for the detection of subnuclear particles in great variety of fields of the science. The use of the technique in measurements of {sup 222} Rn in air have already been established implying better methodologies in the exhibition to the environment until their engraving and reading processes. The SSNTD technique is since a method by comparison since the material it can be used a single time, therefore it requires of calibration in one controlled radon atmosphere, using gauged standards. The objective of this work is to show the calibration of the devices used as radon monitors based on SSNTD. The material used as SSNTD is LR-115 Il. The standardization of the parameters used in the exhibition to radon in air, engraving and reading process, its are based on the response of the LR-115 Il, the one arrangement of the device, engraving speed and mainly the calibration factor. They are considered two types of monitors: Open camera and Closed camera, the difference among the calibration factors of both cameras is the percentage of the descendants of radon in the open camera. The standardized parameters are operation voltage of the counting system; temperature, time and concentration of the engraving solution; and thickness. (Author)

  8. Description of the XXXIV ARCAL Project {sup R}epairing and calibration of electrometers and ionization chambers used in radiotherapy; Descripcion del proyecto ARCAL XXXIV {sup R}eparacion y calibracion de electrometros y camaras de ionizacion utilizados en radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Cruz E, P.; Villaverde L, A. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2002-07-01

    The technological tools from what the humanity has for the illnesses diagnosis and the cancer treatment, are based in great extent in the use of ionizing radiations. This situation worries to the International Atomic Energy Agency (IAEA), which has implemented technical cooperation programs for protecting the human health. In Latin America the ARCAL program (Regional Agreement of Cooperation for Promotion of Nuclear Science and Technology in Latina America and the Caribbean was created. The Project ARCAL XXXIV has as objective to establish three regional centers of repairing, maintenance and electric calibration of clinical dosemeters, equipment made up for an ionization chamber and an electrometer which is used in radiotherapy to generate calibration procedures, personnel training, establishment of an intercomparison net for the electrometers control used as standards and designing current intensity sources which serve as work standards for each one of the participant countries, Mexico is one of them. (Author)

  9. Calibration of the identiFINDER detector for the iodine measurement in thyroid using the Monte Carlo method; Calibracion del detector identiFINDER para la medicion de yodo en tiroides utilizando el metodo Monte Carlo

    Energy Technology Data Exchange (ETDEWEB)

    Ramos M, D.; Yera S, Y.; Lopez B, G. M.; Acosta R, N.; Vergara G, A., E-mail: dayana@cphr.edu.cu [Centro de Proteccion e Higiene de las Radiaciones, Calle 20 No. 4113 e/ 41 y 47, Playa, 10600 La Habana (Cuba)

    2014-08-15

    This work is based on the determination of the detection efficiency of {sup 125}I and {sup 131}I in thyroid of the identiFINDER detector using the Monte Carlo method. The suitability of the calibration method is analyzed, when comparing the results of the direct Monte Carlo method with the corrected, choosing the latter because the differences with the real efficiency stayed below 10%. To simulate the detector their geometric parameters were optimized using a tomographic study, what allowed the uncertainties minimization of the estimates. Finally were obtained the simulations of the detector geometry-point source to find the correction factors to 5 cm, 15 cm and 25 cm, and those corresponding to the detector-simulator arrangement for the method validation and final calculation of the efficiency, demonstrating that in the Monte Carlo method implementation if simulates at a greater distance than the used in the Laboratory measurements an efficiency overestimation can be obtained, while if simulates at a shorter distance this will be underestimated, so should be simulated at the same distance to which will be measured in the reality. Also, is achieved the obtaining of the efficiency curves and minimum detectable activity for the measurement of {sup 131}I and {sup 125}I. In general is achieved the implementation of the Monte Carlo methodology for the identiFINDER calibration with the purpose of estimating the measured activity of iodine in thyroid. This method represents an ideal way to replace the lack of patterns solutions and simulators assuring the capacities of the Internal Contamination Laboratory of the Centro de Proteccion e Higiene de las Radiaciones are always calibrated for the iodine measurement in thyroid. (author)

  10. The CIEMAT Neutron Standards Laboratory (LPN): National Reference Facility for the Calibration of Neutron Measurement Equipment; El Laboratorio de Patrones Neutronicos del CIEMAT: instalacion de referencia nacional para la calibracion de equipos de medida neutronica

    Energy Technology Data Exchange (ETDEWEB)

    Mendez, R.

    2014-04-01

    This is a new facility that, in keeping with the recommendations of standard ISO 8529:1, has 20m5-Be and 252Cf neutron sources stored in water inside a 9 m x 7.5 m x 8 m Irradiation Room with 1.25 m concrete walls that is accessed through a 6.5 ton sliding shielded door. These sources are remotely handled from the Control Room by means of a system that places them in the geometric center of the room and aligns them with the equipment or materials to be irradiated arranged on a base plate. The facility has a series of interlocks that guarantee safety during operations with sources, and the verification measurements have demonstrated the suitability of the shielding used. The LPN will be the answer to the demand for a national laboratory for neutron measurement equipment calibration and equipment and material irradiation under controlled conditions, and it will position our country at the same level as our European partners. (Author)

  11. Control console of the gamma calibration room; Consola de control de la Sala de Calibracion Gamma

    Energy Technology Data Exchange (ETDEWEB)

    Vilchis P, A.E.; Romero G, M. [Instituto Nacional de Investigaciones Nucleares, Ingenieria Electronica, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    1999-07-01

    The Nuclear Centre of Mexico has a Ionizing Radiation Metrology Center (CMRI). This is in charge of the calibration in Mexico and Latin America of equipment dedicated to radiation measurement as industrial, medical as other fields. The importance to ensure that the equipment stay justly calibrated, it is imposed the necessity of automating the different rooms which the CMRI has. in this case it will be exposed the Calibration room for gamma radiation type. The operation of this application was carried out with the LabVIEW development platform and also in C language. The hardware associated is: personal computer with two cards using the 8255 device, 16 channels with optical isolation to manage input/output TTL type, 16 channels with optical isolation to management of charges to 127 V a.c., and 2 channels for 90V d.c. motors. (Author)

  12. Calibration curves for biological dosimetry; Curvas de calibracion para dosimetria biologica

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero C, C.; Brena V, M. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)]. E-mail cgc@nuclear.inin.mx

    2004-07-01

    The generated information by the investigations in different laboratories of the world, included the ININ, in which settles down that certain class of chromosomal leisure it increases in function of the dose and radiation type, has given by result the obtaining of calibrated curves that are applied in the well-known technique as biological dosimetry. In this work is presented a summary of the work made in the laboratory that includes the calibrated curves for gamma radiation of {sup 60} Cobalt and X rays of 250 k Vp, examples of presumed exposure to ionizing radiation, resolved by means of aberration analysis and the corresponding dose estimate through the equations of the respective curves and finally a comparison among the dose calculations in those people affected by the accident of Ciudad Juarez, carried out by the group of Oak Ridge, USA and those obtained in this laboratory. (Author)

  13. Use of the electronic system of portal image for quality control and calibration of the position of blades of multi leaf collimation system of an accelerator Elekta synergy; Uso del sistema electronico de imagen portal para el control de calidad y la calibracion de la posicion de las laminas del sistema de colimacion multilaminas de un acelerador elekta synergy

    Energy Technology Data Exchange (ETDEWEB)

    Sanchez Merino, G.; Cortes Rodicio, J.; Lope Lope, R.; Martin gonzalez, T.; Garcia Fidalgo, M. A.

    2013-07-01

    We present our methodology for quality control of the positioning of the blades using electronic portal imaging system (ILD). Also an example of its usefulness for calibration multi leaf collimation system (MLC). (Author)

  14. Implementation of the method air-kerma product area in KAP camera calibration with reference qualities of X-ray series at the SSDL RQR of Mexico; Implementacion del metodo producto kerma en aire-area en la calibracion de camaras KAP con calidades de referencia de rayos X de la serie RQR en el LSCD de Mexico

    Energy Technology Data Exchange (ETDEWEB)

    Cejudo, Jesus; Tovar, Victor M., E-mail: jesus.cejudo@ini.gob.mx, E-mail: victor.tovar@inin.gob.mx [lnstituto Nacional de lnvestigaciones Nucleares (DMRI/LSDC/lNlN), Ocoyoacac (Mexico). Centro Nuclear Dr. Nabor Carrillo Flores. Laboratorio Secundario de Calibracion Dosimetrica

    2013-10-01

    The X-Ray machines, at a reference laboratory for the instrument calibration in diagnostic radiology, should compliance with the ISO requirements. Sometimes there is not available as much laboratories as needed in Latin American countries. So this project shows the KAP ionization chamber implementation method using the TRS-457 radiation quality from the IAEA at the SSDL of Mexico. The KAP instruments calibration method consists in doing a substitution comparison using a standard reference with traceability to a primary laboratory and a transmission-monitoring chamber that measures the number of photons of the X-ray primary beam. A KAP chamber calibration requires a special array that consists in collocating the chamber in two different positions of its calibration process. Then, with air kerma-area product coefficient together with a corrected electrometer measure at referential conditions, the patient dosimetry magnitudes are calculated. The dosimetry necessity at hospitals always will be in function of possessing a highly reliable calibration coefficient chamber for making these measures. That dosimetry results will help in reducing the total or partial irradiation emitted to the human body of the patient. This is how stochastic risks will be lessened due to diagnostic studies. The purpose of this project is to have a synergy with calibration for making known that the SSDL of Mexico has the technical capacity to act as a link between primary standard dosimetry laboratories and the ionizing radiation equipment users who require that their KAP chamber have traceability from the primary standard to the user. (author)

  15. Geometric calibration of the system of micro-CBCT; Calibracion geometrica de un sistema micro-CBCT

    Energy Technology Data Exchange (ETDEWEB)

    Martinez Rodriguez, I.; Chevalier del Rio, M.; Alieva, T.; Garcia Pinto, D.

    2013-07-01

    The image reconstruction from projections generated with a beam computed tomography system conical (CBCT) requires very precise calibration of a large number of variables that characterize the geometry of the System. Currently, there are several methods of calibration1consistent in the use of specific mannequins. In this work has been used a methodology based on a simple dummy images 2 that has proven results similar or even better compared to other more complex methods. (Author)

  16. Calibration of a PIXE system for analysis of environmental aerosols; Calibracion de un sistema PIXE para analisis de aerosoles ambientales

    Energy Technology Data Exchange (ETDEWEB)

    Aldape, F.; Diaz, R.V.; Anguiano, G.; Flores M, J. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    1991-04-15

    The results of the calibration of the PIXE system for analysis of environmental aerosols in the CNM is presented. The system was calibrated in order to lead to the project study of environmental aerosols using PIXE from the qualitative dimension to the quantitative one. Part of this work it was carried out in collaboration with the expert of the IAEA Dr. D. Crumpton. The calibration of the system it was achieved bombarding thin targets of well-known composition with protons of 2.5 MeV. Using this technique it was determined the absolute elementary concentration in samples of atmospheric aerosols. As an example the calculation of the lead concentration in one of them whose result was of 662 {+-} 49 ng m{sup -3} is presented. This value agrees with the measurements made in the Crocker Nuclear Lab of the UCD. Its are also discussed the reproducibility and precision of the used technique. (Author)

  17. Calibration of a NE213 detector for neutron spectroscopy; Calibracion de un detector de NE213 para espectroscopia de neutrones

    Energy Technology Data Exchange (ETDEWEB)

    Blazquez Martinez, J.; Butragueno Casado, J. L.

    1974-07-01

    This work describes the experimental way followed for getting the calibration of a NE213 detector with a beam of neutrons from the J.E.N. 2 MeV Van de Graaff and using at once pulse shape discrimination. Detector has been used for measuring the spectrum of the fast reactor CORAL-1. There is also included an experimental method in order to get with precision where the Compton edge is placed on the electron spectrum. (Author) 9 refs.

  18. MEPSOCON project: Calibration of Radiometers for High Solar Irradiance; Proyecto MEPSOCON: Calibracion de Radiometros de Alta Irradiancia Solar

    Energy Technology Data Exchange (ETDEWEB)

    Ballestrin, J.; Rodriguez-Alonso, M.

    2006-07-01

    The development of central receiver solar plants is a currently emerging field into renewable energies. For several years various receiver prototypes have been evaluated at the Plataforma Solar de Almeria (PSA). The measurement of the incident solar power on the receiver aperture is fundamental to the estimation of its efficiency. Many factors interfere with this measurement and consequently accuracy is very low. This uncertainty is transmitted to the design of the final solar plant and thereby to its price. The sensors used for this measurement are of small size in comparison with the receiver apertures, therefore different systems are necessary to obtain the incident solar power on the receiver aperture from the individual radiometer measurements. This report presents calibration procedures for the sensor used on the measurement of high solar irradiance and the analysis of the different factors affecting the incident power measurement to significantly reduce its uncertainty. (Author) 16 refs.

  19. Calibration curves for quantifying praseodymium by UV-VIS; Curvas de calibracion para cuantificar praseodimio por UV-VIS

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez M, R.; Lopez G, H.; Rojas H, A. [ININ, 52750 La Marquesa, Estado de Mexico (Mexico)]. e-mail: robertssd1199@yahoo.com.mx

    2007-07-01

    The UV-Vis spectroscopic technique was used to determine the absorption bands depending on the concentration from the praseodymium solutions at pH3. Those more appropriate were in the wavelength of 215 nm, for concentrations of 0.0001-0.026 M, of 481nm, 468 nm and 443 nm, for concentrations of 0.026-0.325 M, and of 589 nm, for concentrations of 0.026-0.65 M of the praseodymium. To these wavelengths the calibration curves were determined, which presented correlation coefficients between 0.9976 and 0.9999 except of the absorption of 589 nm that gave R{sup 2} = 0.9014. (Author)

  20. Calibration of the hydraulic model of a full-scale activated sludge plant; Calibracion hidraulica a escala real de un reactor de lodos activados

    Energy Technology Data Exchange (ETDEWEB)

    Fall, Cheikh [Universidad Autonoma del Estado de Mexico (Mexico); Loaiza-Navia, Jimmy [Servicios de Agua y Drenaje de Monterrey (Mexico)

    2008-04-15

    When planning to simulate a wastewater treatment plant (WWTP) with the activated sludge model number 1 (ASM1), one of the first requirements is to determine the hydraulic model of the reactor. The aim of this study was to evaluate the hydrodynamic regime of the aeration tank of a municipal WWTP by using a rhodamine tracer test and the Aquasim simulation software. A pre-simulation was performed in order to quantify the appropriate colorant mass, set up a sampling plan and evaluate the anticipated visual impact of the tracer test in the river receiving the treated effluents. A tracer test and dynamic flow measurements were carried out, the results of which served to establish and calibrate the hydraulic model. The evaluated tank was physically built as a plug-flow reactor subdivided in 7 compartments, but the study revealed that it is best represented by a model with 5 virtual mixed reactors in series. Through the study, the approach of using a WWTP simulator for hydraulics calibration was shown to be a powerful and flexible tool for designing a tracer test and for identifying adequate tank-in-series models of full-scale activated sludge aeration tanks. [Spanish] Cuando se planea simular una planta de tratamiento con base en el modelo numero 1 de lodos activados (ASM1), uno de los primeros requisitos es determinar el modelo hidraulico del reactor. En este trabajo se estudio el regimen hidrodinamico del tanque de accion de una planta de tratamiento de aguas residuales municipales (PTAR), utilizando una prueba de trazador con rodamina y un programa de simulacion (Aquasim). Se realizo una prueba de trazador con el experimento, lo que permitio determinar la cantidad requerida de trazador, fijar los intervalos de muestreo y limitar el impacto visual anticipado del colorante sobre el rio que recibe el efluente tratado. Se llevaron a cabo la prueba de trazador y la medicion de los perfiles dinamicos de caudales, cuyos resultados sirvieron para establecer y calibrar el

  1. Development of a procedure of calibration of meters of product Kerma-Area; Desarrollo de un procedimiento de calibracion de medidores de producto Kerma-Area

    Energy Technology Data Exchange (ETDEWEB)

    Ginjaume, M.; Roig, M.; Amores, M.; Ortega, X.

    2013-07-01

    The aim of this paper is to present the calibration procedure developed, uncertainties associated and scope of accredited degree. Also the objectives of EURAMET 1177 and calibrations are described conducted by our laboratory in this intercomparison. (Author)

  2. Calibration of film radiochromic EBT2 for sources of I-125 encapsulated; Calibracion de pelicula radiocromica EBT2 para fuentes de I-125 encapsulado

    Energy Technology Data Exchange (ETDEWEB)

    Huerga Cabrerizo, C.; Luquero Llopis, N.; Torre Hernandez, I. de la; Ferrer Garcia, C.; Corredoira silva, E.; Serrada Hierro, A.

    2013-07-01

    This paper determines the calibration curve in absolute dose for sources of I-125 encapsulated to estimate its uncertainty. In order to assess energy dependence is compared with the obtained for an accelerator of 6MV calibration curve. (Author)

  3. Calibration of a Plastic Classification System with the CCW Model; Calibracion de un Sistema de Clasification de Plasticos segun el Modelo CCW

    Energy Technology Data Exchange (ETDEWEB)

    Barcala Riveira, J. M.; Fernandez Marron, J. L.; Alberdi Primicia, J.; Navarrete Marin, J. J.; Oller Gonzalez, J. C.

    2003-07-01

    This document describes the calibration of a plastic Classification system with the CCW model (Classification by Quaternions built Wavelet Coefficients). The method is applied to spectra of plastics usually present in domestic wastes. Obtained results are showed. (Author) 16 refs.

  4. Calibration and quality control of a multi leaf collimator using linear array of detectors; Calibracion y control de calidad de un colimador multilaminas mediante array lineal de detectores

    Energy Technology Data Exchange (ETDEWEB)

    Suero Rodrigo, M. A.; Marques Fraguela, E.

    2011-07-01

    The protocol for calibration and quality control established by Siemens for the multi leaf collimator (MLC) of Primus electron linear accelerators, using the light field coincidence with the beam of radiation to determine the position of the blades. In this paper, we illustrate the use of a calibration method for determining the positions of MLC plates radiologically with the help of a linear array of detectors, based on the proposal Lopes et al (2007).

  5. Evaluation of different calibration curves QA of IMRT plans with radiochromic films; Evaluacion de diversas curvas de calibracion QA de planes de IMRT con peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez Rodriguez, J.; Martin Rincon, C.; Garcia Repiso, S.; Ramos Paheo, J. A.; Verde Velasco, J. M.; Sena Espinel, E. de

    2013-07-01

    The non-linear relationship between dose and the optical density, characteristic plates radiochromic Gafchromic EBT and EBT2, has been studied by various authors, whose publications are proposed different functional forms that fit the specific values measured curves that allow the full range of useful dose calibration. The objective of the work focuses on evaluating the influence of the use of different calibration curves in the dose measurement for quality control of IMRT treatments. (Author)

  6. Monte Carlo simulation: tool for the calibration in analytical determination of radionuclides; Simulacion Monte Carlo: herramienta para la calibracion en determinaciones analiticas de radionucleidos

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez, Jorge A. Carrazana; Ferrera, Eduardo A. Capote; Gomez, Isis M. Fernandez; Castro, Gloria V. Rodriguez; Ricardo, Niury Martinez, E-mail: cphr@cphr.edu.cu [Centro de Proteccion e Higiene de las Radiaciones (CPHR), La Habana (Cuba)

    2013-07-01

    This work shows how is established the traceability of the analytical determinations using this calibration method. Highlights the advantages offered by Monte Carlo simulation for the application of corrections by differences in chemical composition, density and height of the samples analyzed. Likewise, the results obtained by the LVRA in two exercises organized by the International Agency for Atomic Energy (IAEA) are presented. In these exercises (an intercomparison and a proficiency test) all reported analytical results were obtained based on calibrations in efficiency by Monte Carlo simulation using the DETEFF program.

  7. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Cortes P, A.; Garcia D, O.C.; Ortiz P, I.; Becerril V, A. [Instituto nacional de Investigaciones Nucleares, Laboratorio de Patrones Radiactivos, Departamento de Metrologia, A.P. 18-1027. C.P. 11801 Mexico D.F. (Mexico)

    2000-07-01

    With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup 3} contained in polyethylene small bottles of 25 cm{sup 3} elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm{sup 3} of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)

  8. Energy Calibration of a Silicon Detector Using Pure Beta-Emitters; Calibracion Energetica de un Detector de Silicio Mediante Emisores Beta Puros

    Energy Technology Data Exchange (ETDEWEB)

    Borras, C.; Arcos, J. M. los

    1992-07-01

    Energy calibration of SI detectors used in electron spectroscopy 13 commonly performed with conversion electron sources or monoenergetic electrons beams, which are preferred against beta emitters due to the problems arising from their continuous spectra. This paper presents a simple calibration procedure for a PIP-type silicon detector, using 14C, 147Pm, 99{sup T}c and 45Ca sources, that is based on the correspondence between the average channel observed in the experimental spectrum and the mean energy evaluated from the theoretical Fermi distribution for each nuclide. First, a method for evaluating the average channel in the experimental spectrum distorted by the electronic noise is described and its uncertainty estimated. Then, the channel-energy relation ship is established by least squares fitting modified to account for uncertainties in both variables.The calibration has been successfully verified with 147Pm and '09cd sources, showing discrepaneles not greater than 2.5%, within the uncertainties due to the detector resolution and the sources features. (Author)

  9. Characterization of different qualities in X-rays, for instruments calibration in radiological protection; Caracterizacion de diferentes calidades en rayos X, para calibracion de instrumentos en proteccion radiologica

    Energy Technology Data Exchange (ETDEWEB)

    Cejudo A, J.; Tovar M, V. M.; Vergara M, F., E-mail: jesus.cejudo@inin.gob.m [ININ, Departamento de Metrologia de Radiaciones Ionizantes, Laboratorio Secundario de Calibracion Dosimetrica, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)

    2010-09-15

    In the Secondary Laboratory of Dosimetric Calibration in Mexico was realized the qualities characterization of the series X-rays RQR reported in the International Code of Practices in Dosimetry and Diagnostic Radiology No. 457, using attenuator filters of high purity aluminum and ionizing radiation equipment of inherent filtration of 4 mm Be with and emergent X-radiation beam of 40 grades. For the attenuation was used a geometric arrangement with three beam limiters and a monitor camera prepared on the established form in the mentioned technical report, and a spherical ionization chamber with collection volume of 3.6 cm{sup 3}, aligning its geometric center with the focus of X-rays tube to get that the incident radiation direction will be perpendicular to the ionization chamber. From the perspective of the radiological protection is important to know the X-radiation quality for the application dedicated to the instruments calibration and can to give to these the traceability to a reference laboratory, this way the quality combination and reference chamber can give as a result a procedure for the evaluation of the entrance in surface dose to estimate the dose orientate levels, specified in the basic standards of safety. (Author)

  10. Calibration of a plastic detector PVT of large dimensions for the detection gamma radiation; Calibracion de un detector plastico de PVT de grandes dimensiones para la deteccion de radiacion gamma

    Energy Technology Data Exchange (ETDEWEB)

    Armenteros Carmona, J. C.; Suarez navarro, M. J.; Gonzalez-gonzalez, J. A.; Pujol, L.

    2011-07-01

    This article describes the development work on the porch RadSentry Canberra brand that uses 2 plastic scintillation detectors PVT large (172 x 32 x 4 cm). In this paper we present the different calibration necessary for the proper functioning of equipment: channel-energy calibration, energy resolution, and efficiency-energy.

  11. Use of a robotic manipulator in the simulation of the automation of a calibration process of dosemeters; Uso de un manipulador robotico en la simulacion de la automatizacion de un proceso de calibracion de dosimetros

    Energy Technology Data Exchange (ETDEWEB)

    Benitez R, J.S.; Najera H, M.C. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2002-07-01

    The development of a system based in a manipulative robot which simulates the operative sequence in a calibration process of dosemeters is presented. In this process it is performed the monitoring of the dosemeter positions and the calibrator by mean of an arm of articulated robot which develops the movement sequences and the taking a decision based on the information coming from the external sensors. (Author)

  12. Influence of geometry calibration in the determination of gamma emitters in active carbon filters; Influencia de la geometria de calibracion en la determinacion de emisores gamma en filtros de carbon activo

    Energy Technology Data Exchange (ETDEWEB)

    Laurado Tarrago, M.; Tent Petrus, J.

    2013-07-01

    Testing laboratories participate in external quality controls. For these controls are useful it is necessary to eliminate factors that can introduce sources of variability in the results. A factor to consider in this paper as a source of variability is the calibration geometry chosen for the correct quantification of the activity concentrations. This work contains the results of the intercomparison exercise and from them analysed the effect of the geometry of calibration used in the quantification of activity concentrations present in the activated carbon filter. (Author)

  13. Methodology to include a correction for offset in the calibration of a Diode-based 2D verification device; Metodologia para incluir una correccion por offset en la calibracion de un dispositivo de verificacion 2D basado en diodos

    Energy Technology Data Exchange (ETDEWEB)

    Ramirez Ros, J. C.; Pamos Urena, M.; Jerez Sainz, M.; Lobato Munoz, M.; Jodar Lopez, C. A.; Ruiz Lopez, M. a.; Carrasco Rodriguez, J. L.

    2013-07-01

    We propose a methodology to correct doses of device verification 2D MapChek2 planes by offset. This methodology provides an array of correction by Offset applied to the calibration per dose due to the Offset of the diode Central as well as the correction of the Offset of each diode on each acquisition. (Author)

  14. Calibration of a {sup 19} {sup 2}Ir source for high dose brachytherapy using various techniques; Calibracion de una fuente de {sup 192} Ir para braquiterapia de alta tasa de dosis mediante diversas tecnicas

    Energy Technology Data Exchange (ETDEWEB)

    Montilla Prieto, Tedicel C., E-mail: tcdicel@gmaiLcam [Instituto de Oncologia Dr. Miguel Perez Carreno, Barbula (Venezuela, Bolivarian Republic of). Departamento de Fisica y Dosimetria; Padron Rivero, Alvaro D., E-mail: alvarodpadronr@yahoo.com.ve [Universidad de Carabobo, Barbula (Venezuela, Bolivarian Republic of). Facultad de Ciencias de la Salud. Departamento de Ciencias Fisiologicas

    2013-10-01

    In this research we studied three experimental procedures for calibration of a source of {sup 192}Ir to high dose rate for clinical brachytherapy use, and thus were compared and analysis of the advantages and disadvantages of each. For this study we quantified the value of the current kerma rate reference in air by three procedures: source calibration using a well chamber, with an cylindrical ionization chamber in air, and a cylindrical ionization chamber on a phantom, and this magnitude was compared with the value provided by the manufacturer of the source and thereby obtaining the deviation corresponding . Thus, it was found that the deviation corresponding to the source calibration making use of a well chamber, remained within tolerance, while the cylindrical ionization chamber in air and on phantom exceeded the standards established in some documents. However, although both the measurement in air and in the phantom are the procedures for the final calibration source, these can be used to verify that the delivered dose are in tolerance.

  15. Recognition of the dosimetric calibration capacities of Cuba by the International Bureau of Weights and Measures; Reconocimiento de las capacidades de calibracion dosimetrica de Cuba por el Buro Internacional de Pesas y Medidas

    Energy Technology Data Exchange (ETDEWEB)

    Walwyn S, G.; Gutierrez L, S.; Tamayo G, J.A.; Gonzalez R, N.; Alonso V, G. [CPHR, Calle 20 No. 4113 e/ 41 y 47, Playa C.P. 11300, La Habana (Cuba)]. e-mail: gonzalo@cphr.edu.cu

    2006-07-01

    The declared mission of the International Bureau of Weights and Measures are the world uniformity of the measurement, however until some years ago a formal mechanism didn't exist for its complete implementation. With this end arose the Mutual Recognition Agreement whose specific objective is to establish the grade of equivalence of the national standards, the one of mutually recognizing the calibration and measurement certificates and the one of providing to the governments of a sure technical tool in its commercial negotiations and regulatory matters at international level. Cuba like an associated country to the Meter Convention, signed the agreement and it intended to demonstrate the international equivalence of its standards. The best measurement and calibration capacities of the country in the dosimetric magnitudes are in the Secondary Laboratory of Dosimetric Calibration of the Protection and Hygiene of Radiations Center. This capacities were included in the Regional Metrological Organization COOMET in the year 2003. In June of the 2005 the metrological capacities have been approved and published in the databases of the International Bureau of Weights and Measures as demonstration of the high competition of the calibration works that its are carried out in the laboratory. This approval is one of the maximum international recognitions that the patterns of a country can receive and its are the result of 10 years of work of the laboratory like part of the international net OIEA/OMS, which has given it the possibility to gauge the patterns and of adopting internationally validated calibration methodologies. On the other hand, it has been decisive the participation of the laboratory in multiple international comparisons of their patterns, as well as the implementation of a system of administration of the quality credited by the competent national organ. The article reviews the technical work of the laboratory during several years that it gave as result this

  16. Secondary calibration laboratory for dosimetry in levels of therapy at the University of Santiago; Laboratorio secundario de calibracion para dosimetria en niveles de terapia en la Universidad de Santiago

    Energy Technology Data Exchange (ETDEWEB)

    Gomez Rodriguez, F.; Gonzalez Castano, D. M.; Pazos Alvarez, A.

    2011-07-01

    A basic inherent benefits provided by the existence of a traceability chain radiation in any application, add the legal requirement for hospitals as pointed to by the RO. 1566/1998, which sets quality standards in radiotherapy. The decree attributed to hospital specialists radio physics in article 10 the responsibility for determining the acceptance and initial reference state of radiation generating equipment for therapeutic purposes, and the establishment and implementation of quality control programs associated and technical and physical aspects of radiation dosimetry. Different international organizations such as ICRU and IAEA recommendations on maintaining the accuracy of the dose delivered to patients in general, should be placed at least 5% considering the whole chain irradiation. In order to achieve this purpose it is necessary to establish programs of quality control and calibration dosimetric regular basis. The protocol of the IAEA TRS398 recommended dose calibration in water because it is a quantity of interest closest to clinical use and allows a relative uncertainty in the calibration environment reduced to 1%.. (Author)

  17. Use of calibration methodology of gamma cameras for the workers surveillance using a thyroid simulator; Uso de una metodologia de calibracion de camaras gamma para la vigilancia de trabajadores usando un simulador de tiroides

    Energy Technology Data Exchange (ETDEWEB)

    Alfaro, M.; Molina, G.; Vazquez, R.; Garcia, O., E-mail: mercedes.alfaro@inin.gob.m [ININ, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)

    2010-09-15

    In Mexico there are a significant number of nuclear medicine centers in operation. For what the accidents risk related to the transport and manipulation of open sources used in nuclear medicine can exist. The National Institute of Nuclear Research (ININ) has as objective to establish a simple and feasible methodology for the workers surveillance related with the field of the nuclear medicine. This radiological surveillance can also be applied to the public in the event of a radiological accident. To achieve this it intends to use the available equipment s in the nuclear medicine centers, together with the neck-thyroid simulators elaborated by the ININ to calibrate the gamma cameras. The gamma cameras have among their component elements that conform spectrometric systems like the employees in the evaluation of the internal incorporation for direct measurements, reason why, besides their use for diagnostic for image, they can be calibrated with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity distributed homogeneously in the human body, or located in specific organs. Inside the project IAEA-ARCAL-RLA/9/049-LXXVIII -Procedures harmonization of internal dosimetry- where 9 countries intervened (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). It was developed a protocol of cameras gamma calibration for the determination in vivo of radionuclides. The protocol is the base to establish and integrated network in Latin America to attend in response to emergencies, using nuclear medicine centers of public hospitals of the region. The objective is to achieve the appropriate radiological protection of the workers, essential for the sure and acceptable radiation use, the radioactive materials and the nuclear energy. (Author)

  18. System for calibration of instruments of x-ray measurement (CIR-X) applying the PGCS; Sistema para calibracion de instrumentos de medicion de rayos X (CIR-X) aplicando el PGCS

    Energy Technology Data Exchange (ETDEWEB)

    Gaytan G, E.; Rivero G, T.; Cruz E, P.; Tovar M, V.M.; Vergara M, F.J. [ININ, 52750 La Marquesa, Estado de Mexico (Mexico)]. e-mail: egg@nuclear.inin.mx

    2007-07-01

    The Department of Metrology of Ionizing Radiations of the ININ carries out calibration of instruments for X-ray measurement that determine the operation parameters in X-ray diagnostic machines of the health and private sectors. To facilitate this task, the Department of Automation and Instrumentation developed a system for acquisition and signals processing coming from a reference voltage divider with traceability at NIST that is connected directly to the X-rays tube. The system is integrated by the X-ray unit, the X-ray measurement equipment Dynalizer IIIU of RADCAL, a data acquisition card, a personal computer and the acquisition software and signals processing. (Author)

  19. Proposal of a calibration protocol of gamma chambers for estimation of the radionuclides incorporation in emergency situations; Propuesta de un protocolo de calibracion de camaras gamma para estimacion de la incorporacion de radionucleidos en situaciones de emergencia

    Energy Technology Data Exchange (ETDEWEB)

    Dantas, B.M.; Lucena, E.; Dantas, A.L.A.; Araujo, F.; Melo, D. [Instituto de Radioprotecao e Dosimetria, Rio de Janeiro Brasil (Brazil); Teran, M.; Paolino, A. [Facultad de Quimica, Montevideo (Uruguay); Hermida, J.C. [Hospital de Clinicas, Facultad de Medicina, Montevideo (Uruguay); Rojo, A. [Autoridad Regulatoria Nuclear, Buenos Aires (Argentina); Puerta, J.A.; Morales, J. [Universidad Nacional de Colombia, Medellin (Colombia); Lopez B, G.M. [Centro de Proteccion e Higiene de las Radiaciones, Ciudad de la Habana (Cuba); Alfaro, M.; Ruiz, M.A. [Instituto Nacional de Investigaciones Nucleares (Mexico); Videla, R.; Pinones, O. [Comision Chilena de Energia Nuclear, Santiago (Chile); Gonzalez, S. [Instituto Peruano de Energia Nuclear, Lima (Peru); Navarro, T. [Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas, Madrid (Spain); Cruz S, R. [International Atomic Energy Agency, Vienna (Australia)

    2006-07-01

    In the last years in several countries has come increasing the concern with the possibility of accidents related to the transport and manipulation of open sources used in nuclear medicine. This carried out to the search of alternative methods for the monitoring of workers and individuals of the public exposed to the radionuclides incorporation like {sup 131} I, {sup 201} Tl, {sup 153} Sm among others. One of the options to assist the demand for monitoring of the radionuclides incorporation is the use of gamma chambers that are medical diagnostic equipment available in the own centers of nuclear medicine. The gamma chambers are used to obtain images of patient to which are administered a radionuclide of well-known activity with diagnostic purposes. These equipment have among its components elements that spectrometric systems like those used in the evaluation of the internal incorporation for direct measurements, reason why besides its use for diagnosis by image they can be gauged with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity, distributed homogeneously in the human body or located in specific organs. This work presents the development of a calibration protocol of gamma chambers for the in vivo determination of radionuclides and it proposes the implementation of the protocol in centers of nuclear medicine of the 9 countries participants of the project OAS-ARCAL-RLA/9/049-LXXVIII - Harmonization of procedures of internal dosimetry (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). The protocol is the base to establish an integrated net to attend in the response to emergencies using nuclear medicine centers of public hospitals of the region. The proposal is an additional alternative for the monitoring of people in emergency situations where it is possible and feasible the use of the gamma chambers. This would avoid the person's transport with incorporation suspicion for

  20. Calibration of the High Resolution Gamma Spectrometry Detection Systems from the Environmental Radioactivity Unit and Radiological Surveillance (ARU and RV); Calibracion de los Sistemas Detectores de Espectrometria Gamma de Alta Resolucion de la Unidad de Radiactividad Ambiental y Vigilancia Radiologica (URA y VR)

    Energy Technology Data Exchange (ETDEWEB)

    Suarez-Navarro, J. A.; Gasco, C.; Arana, M.; Suanez, A.

    2012-11-01

    In accordance with international regulations and with the quality criteria imposed by the internal quality system of the laboratory of gamma spectrometry, which belongs to the Unit of Environmental Radioactivity and Radiation Monitoring (URA y VR), it is necessary to perform a periodic calibration of the measurement systems of gamma spectrometry. The equipment calibration carries with the need for the preparation of the geometries of measurement and calibration in energy and in efficiencies. This report reflects the procedures for the preparation of the wide range of geometries available in the laboratory, as well as the different calibrations performed with those geometries in almost all the laboratory gamma detectors using the CANBERRA software Genie 2k. (Author) 9 refs.

  1. Calculation of dose absorbed for the verification of the calibration curves UH-Density electronic relative obtained with various mannequins; Calculo de dosis absorbida para la verificacion de las curvas de calibracion UH-Densidad electronica relativa obtenidas con distintos maniquies

    Energy Technology Data Exchange (ETDEWEB)

    Perez Alvarez, M. E.; Sena Espinel, E. de; Delgado Aparicio, J. M.; Martin Rincon, C.; Garcia Repiso, S.; Ramos Pacho, J. A.; Verde Velasco, J. M.; Gomez Gonzalez, N.; Cons Perez, N.; Saez Beltran, M.

    2013-07-01

    In order to discern with what calibration curve is obtained a more accurate dosimetry distribution, dose measurements are carried out on the treatment unit and the values found are evaluated. (Author)

  2. Calibration of 232U solution; Calibracion de una disolucion de 232U

    Energy Technology Data Exchange (ETDEWEB)

    Galan Valera, M. P.; Acena Berrenechea, M. L.

    1988-07-01

    A method for as certain the activity by alpha spectroscopy with semiconductor detectors, of a solution of 232U is presented. It consists of the comparison with a 233U solution activity previously measured in a gridded ionization chamber of 2{pi} geometry. The total measurement uncertainty is about + - 0,02. (Author) 9 refs.

  3. Flow measurement in a 170-MW hydraulic turbine using the Gibson method; Medicion del flujo de una turbina hidraulica de 170 MW utilizando el metodo Gibson

    Energy Technology Data Exchange (ETDEWEB)

    Urquiza, Gustavo [Universidad Autonoma del Estado de Morelos (Mexico); Adamkowski, Adam [The Szewalski Institute of Fluid-Flow Machinery (Poland); Kubiak, Janusz; Sierra, Fernando [Universidad Autonoma del Estado de Morelos (Mexico); Janicki, Waldemar [The Szewalski Institute of Fluid-Flow Machinery (Poland); Fernandez, J. Manuel [Comision Federal de Electricidad (Mexico)

    2007-07-15

    This paper describes the methodology applied for measuring water flow through a 170-MW hydraulic turbine. The flow rate was measured using the pressure-time method, also known as the Gibson method. This method uses the well-known water hammer phenomenon in pipelines; in turbine penstocks, for instance. The version of this method used here is based on measuring, during total stop of the water stream, the time-history of pressure change in one section of the turbine penstock and relate it to the pressure in the upper reservoir to which the penstock is connected. The volumetric flow rate is determined from the relevant integration of the measured temporary pressure rise. Flow measurement was possible this way because the influence of the penstock inlet was negligible as far as an error of the measurement is concerned. The length of the penstock was 300 m. Previous experience and a standard IEC-41-1991 were the criteria adopted and applied. A fast and efficient acquisition system, including a 16 bit card, was used. The flow rate was calculated using a computer program developed and tested on several cases. The results obtained with the Gibson method were used for calibration of the on-line flow measuring system based on the Winter-Kennedy method as one of the index methods. This method is very often used for continuous monitoring of the flow rate through hydraulic turbines, when the calibration has been done on site by using the results of measurements obtained by the absolute method. Having measured the flow rate and output power, the efficiency was calculated for any operating conditions. A curve showing the best operating conditions based on the highest efficiency is presented and discussed. The details of the instrumentation, its installation, and the results obtained are discussed in the paper. [Spanish] Este articulo describe la metodologia aplicada para la medicion del flujo en una turbina hidraulica de 170 MW. El flujo se midio utilizando el metodo de presion

  4. Calibration and study of the measurement capabilities of real-time gamma spectrometry equipment developed for the renewal of the network of stations automatic monitoring environmental radiation of the Generalitat of Catalunya; Calibracion y estudio de las capacidades de medida de los equipos de espectrometria gamma en tiempo real desarrollados para la renovacion de la Red de Estaciones Automaticas de vigilancia radiologia ambiental de la Generalitat de Catalunya

    Energy Technology Data Exchange (ETDEWEB)

    Casanovas, R.; Prieto, E.; Salvado, M.

    2014-10-01

    The renewal of the automatic environmental radiation surveillance network of the Generalitat of Catalunya has been carried out through the development and implementation of gamma-spectrometry-based equipment. the monitors use scintillation crystals, either Na(TI) or LaBr{sub 3}(Ce) and currently, there are 3 types of equipment water radioactivity monitors, aerosols on a particulate filter monitors and direct measurement monitors. In this paper, we expose the basic features of its operation, the details of their calibration and the minimum detectable activity concentration for some isotopes. (Author)

  5. Paleoclimatología y cronología del Würm reciente: un intento de síntesis

    Directory of Open Access Journals (Sweden)

    Ángel RIVERA ARRIZABALAGA

    2009-12-01

    Full Text Available RESUMEN: Las oscilaciones climáticas observadas en los núcleos polares (GRIP, GISP2 y la escala temporal que forman, pueden servirnos como marco de referencia general de las variaciones climáticas representadas en los yacimientos arqueológicos. De igual modo, los registros de núcleos marinos (MD95- 2042 y MD95-2043, junto con los terrestres del lago Grande de Monticchio y las dataciones U/Th de los espeleotemas de la cueva de Villars (SW de Francia, nos permiten comprobar la repercusión de estos cambios en nuestro continente. En su desarrollo, se hace evidente la necesidad de realizar calibraciones sobre las dataciones realizadas por el método 14C, lo que actualmente se está intentado por medio de síntesis sobre los datos recogidos en diversos núcleos sedimentarios, donde es posible conocer la datación de sus niveles por medios diferentes a los radiocarbónicos.ABSTRACT: Climatic oscillations determined in polar cores (GRIP, GISP2 and the chronological framework modeled in them can be used as sets of global references for determine climate oscillations represented in archaeological deposits. Record of marine cores (MD95-2042 and MD95-2043, with the terrestrial one of the Big Lake of Monticchio and the U/Th dates of Villars's cave stalactite (SW France, allow us to verify the impact of these changes in Europe. 14C dates need to be calibrated. In order to obtain a reliable calibration of them, nowadays it is in progress the build of framework synthesizing information gathered in those sedimentary cores containing levels that offer more dating possibilities that only radiocarbon methodology.

  6. Utilidad del dispositivo portátil PIKO6 para la detección precoz de EPOC en Atención Primaria

    OpenAIRE

    Hidalgo Sierra, Vanesa

    2014-01-01

    [ES]Partiendo de la premisa que la EPOC es una enfermedad muy prevalente, pero infradiagnosticada, el presente trabajo ha consistido en la validación del dispositivo portátil Piko-6 para la detección precoz de EPOC en Atención Primaria, dado que el dispositivo portátil Piko-6 es mucho más barato, portátil y manejable, y no requiere adiestramiento especial por el personal sanitario ni de calibraciones periódicas. El objetivo principal de nuestro estudio fue validar la hipótesis de trabajo d...

  7. Influencia del momento de inercia del tambor y de diferentes ángulos de alimentación constantes sobre el calibre de las partículas de forraje procesado con picadores del tipo de tambor con alimentación manual. Parte I

    Directory of Open Access Journals (Sweden)

    Pedro A. Valdés Hernández

    2010-01-01

    Full Text Available La investigación realizada, presenta como objetivo determinar la influencia del momento de inercia del tambor picador y de diferentes ángulos de alimentación constantes, sobre el calibre de las partículas de forraje procesado con máquinas picadoras de forraje de tallos gruesos del tipo de tambor con alimentación manual. En la investigación se aplica la modelación físico-matemática del proceso tecnológico, cuyos resultados se toman como base para el diseño de experimentos realizados durante el procesamiento de tallos de la variedad de Caña de Azúcar C323-68, siendo la de mayor uso en Cuba para la alimentación del ganado. La modelación físico matemática del proceso permite la predicción del calibre de las partículas con errores no superiores a 15,3%, con un nivel de significación del 1% para valores racionales del momento de inercia del tambor, calculado sobre bases teóricas. La racionalización del momento de inercia en la picadora de forraje de producción nacional MF-IIMA modelo EM-01 produjo un incremento en la calidad del desmenuzado, que se manifestó en un mayor porcentaje de partículas con calibres por debajo de 20 mm entre 43 y 48%. La modelación teórica resultó una valiosa herramienta de trabajo, posibilitando el perfeccionamiento del diseño de este tipo de picadora de forraje.

  8. Responsabilidad social: una ruta para alcanzar la excelencia en laboratorios de ensayos y calibraciones

    OpenAIRE

    Herrera-Basurto, Raúl; Mercader-Trejo, F.; Torres Rivera, O.

    2013-01-01

    Se propone un modelo de gestión estratégico para incluir la gestión de la Responsabilidad Social (RS) como una herramienta que permita no sólo la responsabilidad por la calidad de los resultados y el compromiso con los clientes, la sociedad y los grupos de interés, sino también como una herramienta para lograr la excelencia de la organización. Una nueva concepción del laboratorio es una "condición absoluta" para su legitimidad hoy y mañana. El modelo de gestión es integrador y compatible, don...

  9. Calibration in water films GAFCHROMIC EBT radiochromic-2. Effects of Drying; Calibracion en agua de las peliculas radiocromicas Gafchromic EBT-2. Efectos de Secado

    Energy Technology Data Exchange (ETDEWEB)

    Herrero, C.; Perez-Alija, J.; Alaman, C.; Almansa, J.; Vilches, M.

    2011-07-01

    Recent studies [1] show that immersion in water GAFCHROMIC EBT radiochromic film-2 has on these two effects: a progressive diffusion of water across its borders (fact which tells us the manufacturer and the effect of which we can avoid choosing a proper analysis region) and a uniform moisture through the polyester film, producing a variation in the measurement of optical density (OD) of the film. The latter effect is negligible for immersion times of less than 30 minutes. We study the effect of hydration and drying in these films when they have been submerged for 24 hours.

  10. Characterization and use of shadow cone in a calibration laboratory for neutron; Caracterizacion y empleo de conos de sombra en un laboratorio para calibracion neutronica

    Energy Technology Data Exchange (ETDEWEB)

    Gallego Diaz, E.; Valle Diez, A. del; Lorente Fillol, A.; Vega Carrillo, H. R.

    2011-07-01

    We present the design, characterization and use of shadow cones neutron measurements of the Laboratory of Nuclear Engineering Department of the ETSII-UPM (LMN-UPM), with an irradiation room size 9 mx 16 mx 8 m in that used an {sup 2}41Am-Be source of 11 GBq (3 Ci).

  11. Calibración y Validación del Modelo SIMBA-POP en la Zona Bananera del Urabá Antioqueño, Colombia SIMBA-POP Model: Calibration and Validation in Urabá - Antioquia, Colombia

    Directory of Open Access Journals (Sweden)

    Germán Wbeimar Guarín Giraldo

    2012-06-01

    Full Text Available Resumen. SIMBA-POP es un modelo desarrollado en Martinica (Francia, basado en los conceptos de cohortes y cadenas de cohortes de plantas, simula la dinámica poblacional de plantaciones de banano. El modelo original, modificado por Ochoa (2010, se programó en el software EVOLUCIÓN 4.0, desarrollado por el grupo SIMON de la Universidad Industrial de Santander y se calibró y validó para las condiciones de temperatura media habituales en el Urabá Antioqueño; la más importante zona productora de banano en Colombia. En el proceso de calibración se ajustan los ciclos de embolse y cosecha, así como los tiempos en que ocurren los máximos valores de embolse. Para controlar la calibración se usó la desviación media (DM, el error medio de predicción (EMP, el error cuadrático medio de predicción (ECMP, la raíz del error cuadrático medio (RECMP, curvas de masa y coeficiente de correlación lineal. Los resultados mostraron que SIMBA-POP es capaz de simular de forma verosímil la dinámica del ciclo anual de producción en Urabá; particularmente reproduce bien los tiempos de los picos y los valles de producción, aunque la magnitud de estos valles y picos es altamente sensible a los parámetros del modelo y condiciones climáticas no consideradas.Abstract. SIMBA - POP is a model developed in Martinique (France, which uses the concepts of cohorts and cohort chains to simulate the population dynamics of a commercial plantation of bananas. The original model was slightly modified by Ochoa (2010 in order to correct for an error in the flowering rate curve. The model was programmed in EVOLUCION 4.0. SIMBA - POP was then calibrated and validated for the normal climatic pattern in Urabá (Antioquia. Care was taken that the model adequately represent bagging and harvesting cycles, and the time at which maximum values of bagging (flowering were reached. To evaluate the calibration we used average deviation (DM, the average error of prediction (EMP

  12. Soil Moisture Measurement through Time Domain Reflectometry (TDR). Irrigation Application; Medicion de la humedad del suelo por reflectometria en el dominio temporal (TDR). Aplicacion en irrigacion

    Energy Technology Data Exchange (ETDEWEB)

    Quinonez Pedroza, Hector E.; Ruelle, Pierre; Nemeth Ildiko [Cemagre, (France)

    2002-03-01

    Knowledge of how moisture varies in the soil is essential to determine the actual rate of evapotranspiration in crops and, thus, the amount of water necessary for irrigation. Lisimeters, tensiometers and the neutron probe have been widely used for precise measurement of soil moisture in order to calibrate models of irrigation scheduling or to calculate volumes of irrigation water. Despite the usefulness of such measurements, these methods have limited use because of problems like high cost of lisimeters and neutron probes as well as their strict regulation, limited range of validity of tensiometers. Moreover, the measurements obtained from these methods generally describe noncontinuous points in time. Time Domain Reflectometry (TDR) is a viable to such methods because of its precision and the continuity of its measurements. This study evaluates soil moisture, using the neutron and TDR methods in maize crop irrigated by gravity in closed furrows. The results show that even though the absolute determinations of moisture are different in the 30-90 cm profiles, moisture variations were similar in all cases and respond consequently to the additions of water and to evapotranspiration. Neutron probe drawback become evident on its no-continuous measurements contrasting with the continuous measurements of TDR, which allows a detailed analysis of the infiltration and evapotranspiration phenomenal at instantaneous time scales. [Spanish] El conocimiento de la variacion de la humedad del suelo es esencial para determinar la evapotranspiracion real de los cultivos. Esta ultima, a su vez, permite conocer las cantidades necesarias de agua de riego. Los lisimetros, los tensiometros y el aspersor de nuestro s han sido ampliamente usados para la medicion precisa de la humedad del suelo con propositos de calibracion de modelos de programacion del riego o para su utilizacion directa en el calculo de volumenes de riego. A pesar de la importancia de tales mediciones, su uso es muy

  13. Use of Near Infrared Reflectance (NIR Spectroscopy to Predict Chemical Composition of Forages in Broad-Based Calibration Models Uso de la Espectroscopía de Reflectancia en el Infrarrojo Cercano (NIR Para Predecir la Composición Química de Forrajes en Modelos de Calibración Amplia

    Directory of Open Access Journals (Sweden)

    Jaime Garcia

    2006-03-01

    Full Text Available The objective of the study was to evaluate the potential of near infrared reflectance (NIR spectroscopy as a rapid method to predict the chemical composition of forage in broad-based calibration models. In total, 650 samples representing a wide range of chemical characteristics, phenological states and origins were scanned in an NIR instrument. The coefficient of determination in calibration (R² and standard error in cross validation (SECV for the NIR calibration models were as follows: dry matter 0.95 (SECV: 0.7%, crude protein 0.98 (SECV: 0.98%, ash 0.90 (SECV: 0.99%, in vitro organic matter digestibility 0.90 (SECV: 3.6%, acid detergent fiber 0.95 (SECV: 2.0% and neutral detergent fiber 0.86 (SECV: 5.4% on a dry matter basis. The results demonstrated the potential of NIR to predict the chemical composition of different forage plant species ; however, it is suggested that the technique could be used as a routine procedure to apply in breeding programs only if calibration is done for each species, season and particular conditions.El objetivo del trabajo fue evaluar el potencial de la espectroscopía de reflectancia en el infrarrojo cercano (NIR como método rápido para la predicción de la composición química de forrajes en modelos de calibración amplia. Un total de 650 muestras de forrajes de un amplio rango de características químicas, origen y estados fenólogicos, fueron analizadas en un instrumento NIR en reflectancia. Los coeficientes de determinación en calibración (R² y los errores estándares de la validación cruzada (SECV para las calibraciones NIR fueron 0,95 (SECV: 0,7%, 0,98 (SECV: 0,98%, 0,90 (SECV: 0,99%, 0,95 (SECV: 2,0%, 0,86 (SECV: 5,4% y 0,90 (SECV: 3,6% para materia seca (MS, proteína cruda (PC, cenizas, fibra detergente ácido (FDA, fibra detergente neutro (FDN, y digestibilidad in vitro de la materia orgánica (DIVMO, base materia seca, respectivamente. Los resultados demostraron el potencial del NIR para

  14. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...... uncertainty was verified from independent measurements of the same sample by demonstrating statistical control of analytical results and the absence of bias. The proposed method takes into account uncertainties of the measurement, as well as of the amount of calibrant. It is applicable to all types of...

  15. Camera calibration

    OpenAIRE

    Andrade-Cetto, J.

    2001-01-01

    This report is a tutorial on pattern based camera calibration for computer vision. The methods presented here allow for the computation of the intrinsic and extrinsic parameters of a camera. These methods are widely available in the literature, and they are only summarized here as an easy and comprehensive reference for researchers at the Institute and their collaborators.

  16. Problema de calibración de mercado y estructura implícita del modelo de bonos de Black-Cox || Market Calibration Problem and the Implied Structure of the Black-Cox Bond Model

    Directory of Open Access Journals (Sweden)

    Santana Jiménez, Lisette Josefina

    2010-01-01

    Full Text Available El principal resultado de este artículo consiste en la resolución del problema inverso del modelo de Black-Cox (1976, usando el método propuesto por Sukhomlin (2007. Se parte del enfoque retrógrado (backward para obtener una expresión exacta de la volatilidad implícita en función de parámetros cuantificables con datos de mercado y de variables conocidas. Se descubre la existencia de dos valores de la volatilidad para un activo subyacente en elmodelo referido, lo que indica que las asunciones tradicionales no lo definen de manera unívoca. Se encuentra la causa de que el modelo de Black-Cox contenga dos valores de la volatilidad. Además, se lleva a cabo una simulación, afin de verificar, numéricamente, que la expresión obtenida para la volatilidad es la inversión de la fórmula que representa la probabilidad de que la firma no alcance un nivel de insolvencia antes del tiempo de madurez de la deuda. Finalmente, se resuelve el problema de calibración de mercadodesde el punto de vista directo (forward, encontrándose una expresión que resulta de mayor utilidad para los agentes de mercado. || The main result of this paper consists in the resolution of the inverse problem for the Black-Cox (1976 model, using the method proposed by Sukhomlin (2007. Based on the backward approach, we obtain an exact expression of the implied volatility expressed as a function of quantifiable market parameters and known variables. We discover the existence of two values of the volatility for an underlying asset, in the referred model, which means that the model's traditional assumptions do not define it univocally. We findthe cause that the Black-Cox model contains two values of the volatility. Besides, we carry out a simulation in order to verify, numerically, that our volatility expression is in fact the inversion of the formula that represents the probability that the firm has not reached the reorganization boundary before the debt expires. Finally, we

  17. Calibration Binaries

    Science.gov (United States)

    Drummond, J.

    2011-09-01

    Two Excel Spreadsheet files are offered to help calibrate telescope or camera image scale and orientation with binary stars for any time. One is a personally selected list of fixed position binaries and binaries with well-determined orbits, and the other contains all binaries with published orbits. Both are derived from the web site of the Washington Double Star Library. The spreadsheets give the position angle and separation of the binaries for any entered time by taking advantage of Excel's built in iteration function to solve Kepler's transcendental equation.

  18. ALTEA calibration

    Science.gov (United States)

    Zaconte, V.; Altea Team

    The ALTEA project is aimed at studying the possible functional damages to the Central Nervous System (CNS) due to particle radiation in space environment. The project is an international and multi-disciplinary collaboration. The ALTEA facility is an helmet-shaped device that will study concurrently the passage of cosmic radiation through the brain, the functional status of the visual system and the electrophysiological dynamics of the cortical activity. The basic instrumentation is composed by six active particle telescopes, one ElectroEncephaloGraph (EEG), a visual stimulator and a pushbutton. The telescopes are able to detect the passage of each particle measuring its energy, trajectory and released energy into the brain and identifying nuclear species. The EEG and the Visual Stimulator are able to measure the functional status of the visual system, the cortical electrophysiological activity, and to look for a correlation between incident particles, brain activity and Light Flash perceptions. These basic instruments can be used separately or in any combination, permitting several different experiments. ALTEA is scheduled to fly in the International Space Station (ISS) in November, 15th 2004. In this paper the calibration of the Flight Model of the silicon telescopes (Silicon Detector Units - SDUs) will be shown. These measures have been taken at the GSI heavy ion accelerator in Darmstadt. First calibration has been taken out in November 2003 on the SDU-FM1 using C nuclei at different energies: 100, 150, 400 and 600 Mev/n. We performed a complete beam scan of the SDU-FM1 to check functionality and homogeneity of all strips of silicon detector planes, for each beam energy we collected data to achieve good statistics and finally we put two different thickness of Aluminium and Plexiglas in front of the detector in order to study fragmentations. This test has been carried out with a Test Equipment to simulate the Digital Acquisition Unit (DAU). We are scheduled to

  19. Trinocular Calibration Method Based on Binocular Calibration

    OpenAIRE

    CAO Dan-Dan; Luo, Chun; GAO Shu-Yuan; Wang, Yun; Li, Wen-Bin; XU Zhen-Ying

    2012-01-01

    In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error) of the global calibration of ...

  20. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina

    2016-05-02

    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  1. Calibration curve to establish the exposure dose at Co{sup 60} gamma radiation; Curva de calibracion para establecer dosis de exposicion a radiacion gamma de Co{sup 60}

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero C, C.; Brena V, M. [Departamento de Biologia, Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, Mexico D.F. (Mexico)

    2000-07-01

    The biological dosimetry is an adequate method for the dose determination in cases of overexposure to ionizing radiation or doubt of the dose obtained by physical methods. It is based in the aberrations analysis produced in the chromosomes. The behavior of leisure in chromosomes is of dose-response type and it has been generated curves in distinct laboratories. Next is presented the curve for gamma radiation produced in the National Institute of Nuclear Research (ININ) laboratory. (Author)

  2. Radiological protection and calibration of an activity meter with cesium and barium sources in a nuclear medicine center; Proteccion radiologica y calibracion de un activimetro con fuentes de cesio y bario en un centro de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Morales L, M.E. [IPEN INEN Centro de Medicina Nuclear, Av. Aviacion 3799, Lima 34 (Peru)

    2005-07-01

    Presently work is shown the results when gauging a team Deluxe Isotope (Caliper 11) with some sources of Cesium 137 and Barium 133, in a Center of Medicine Nuclear that operates from the anus 1983 in a modern building inside the one Institute of Illnesses Neoplasia (Inn). This Center was equipped initially with teams donated by the International Organism of Atomic Energy (Oa) with those that it develops assistance, educational works and of investigation, giving services to patient of the Inn and other public and private medical centers. (Author)

  3. Trinocular Calibration Method Based on Binocular Calibration

    Directory of Open Access Journals (Sweden)

    CAO Dan-Dan

    2012-10-01

    Full Text Available In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error of the global calibration of the two camera pairs in the experiments can be as low as 0.277% and 0.328% respectively. Experiment results show that this method is feasible, simple and effective, and has high precision.

  4. Automated Camera Calibration

    Science.gov (United States)

    Chen, Siqi; Cheng, Yang; Willson, Reg

    2006-01-01

    Automated Camera Calibration (ACAL) is a computer program that automates the generation of calibration data for camera models used in machine vision systems. Machine vision camera models describe the mapping between points in three-dimensional (3D) space in front of the camera and the corresponding points in two-dimensional (2D) space in the camera s image. Calibrating a camera model requires a set of calibration data containing known 3D-to-2D point correspondences for the given camera system. Generating calibration data typically involves taking images of a calibration target where the 3D locations of the target s fiducial marks are known, and then measuring the 2D locations of the fiducial marks in the images. ACAL automates the analysis of calibration target images and greatly speeds the overall calibration process.

  5. ORNL calibrations facility

    International Nuclear Information System (INIS)

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  6. Analytical multicollimator camera calibration

    Science.gov (United States)

    Tayman, W.P.

    1978-01-01

    Calibration with the U.S. Geological survey multicollimator determines the calibrated focal length, the point of symmetry, the radial distortion referred to the point of symmetry, and the asymmetric characteristiecs of the camera lens. For this project, two cameras were calibrated, a Zeiss RMK A 15/23 and a Wild RC 8. Four test exposures were made with each camera. Results are tabulated for each exposure and averaged for each set. Copies of the standard USGS calibration reports are included. ?? 1978.

  7. Spiral reader calibration

    International Nuclear Information System (INIS)

    The method to calibrate the spiral reader (SR) is presented. A brief description of the main procedures of the calibration program SCALP, adapted for the IHEP equipment and purposes, is described. The precision characteristics of the IHEP SR have been analysed on the results, presented in the form of diagrams. There is a calibration manual for the user

  8. Residual gas analyzer calibration

    Science.gov (United States)

    Lilienkamp, R. H.

    1972-01-01

    A technique which employs known gas mixtures to calibrate the residual gas analyzer (RGA) is described. The mass spectra from the RGA are recorded for each gas mixture. This mass spectra data and the mixture composition data each form a matrix. From the two matrices the calibration matrix may be computed. The matrix mathematics requires the number of calibration gas mixtures be equal to or greater than the number of gases included in the calibration. This technique was evaluated using a mathematical model of an RGA to generate the mass spectra. This model included shot noise errors in the mass spectra. Errors in the gas concentrations were also included in the valuation. The effects of these errors was studied by varying their magnitudes and comparing the resulting calibrations. Several methods of evaluating an actual calibration are presented. The effects of the number of gases in then, the composition of the calibration mixture, and the number of mixtures used are discussed.

  9. Separation of 210Pb, 210Bi and 210Po by ion exchange and their Iiquid scintillation standardization; Separacion del 210Pb, 210Bi y 2I0Po mediante columna de cambio ionico y su calibracion por centelleo liquido

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez, L.; Jimenez, A.; Grau, A.

    1996-07-01

    We applied the CIEMAT/NIST method and alpha/beta discrimination to ''210Pb samples in equilibrium with its daughters, by preparing homogeneous and gel samples. The stability of samples was tested in different available cocktails, HiSafe''TM II, HiSafe''TM III, Ultima-Gold''TM, Ultima-Gold''TM XR, Ultima-Gold''TM AB, Insta-Gel''R and e Insta-Gel''R lI. Also we analyzed the disequilibrium of the radioactive chain 210Pb+210Bi+210Po, achieving an excellent agreement between the results of the spectrum unfolding method and the experimental values. (Author) 13 refs.

  10. Preparation and LSC Standardization of ''89Sr (DNP) Using the CIEMAT/NIST Method; Preparacion del ''89Sr(DNP) y calibracion por centelleo liquido, mediante el metodo CIEMAT/NIST

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Barquero, L.; Los Arcos Merino, J. M.; Grau Malonda, A.

    1994-07-01

    A procedure for preparation of liquid scintillation counting samples of the strontium DNP complex, labelled with ''89Sr, is described, the chemical quench, the counting stability and spectral evolution of this compound is studied in six scintillators, Toluene, Toluene-alcohol, Dioxane-naphthalene, HiSafe II, Ultima- Gold and Instagel. The liquid scintillation standardization of 89Sr-DNP by the CIEMAT/NIST method, using HiSafe II and Ultima-Gold scintillators, has been carried out. The discrepancies between experimental and computed efficiencies are lower than 0.38% and 0.48%, respectively. The solution has been standardized in terms of activity concentration to an overall uncertainty of 0,38%. (Author) 10 refs.

  11. RF impedance measurement calibration

    International Nuclear Information System (INIS)

    The intent of this note is not to explain all of the available calibration methods in detail. Instead, we will focus on the calibration methods of interest for RF impedance coupling measurements and attempt to explain: (1). The standards and measurements necessary for the various calibration techniques. (2). The advantages and disadvantages of each technique. (3). The mathematical manipulations that need to be applied to the measured standards and devices. (4). An outline of the steps needed for writing a calibration routine that operated from a remote computer. For further details of the various techniques presented in this note, the reader should consult the references

  12. The COS Calibration Pipeline

    Science.gov (United States)

    Hodge, Philip E.; Keyes, C.; Kaiser, M.

    2007-12-01

    The COS calibration pipeline (CALCOS) includes three main components: basic calibration, wavelength calibration, and spectral extraction. Calibration of modes using the far ultraviolet (FUV) and near ultraviolet (NUV) detectors share a common structure, although the individual reference files differ and there are some additional steps for the FUV channel. The pipeline is designed to calibrate data acquired in either ACCUM or time-tag mode. The basic calibration includes pulse-height filtering and geometric correction for FUV, and flat-field, deadtime, and Doppler correction for both detectors. Wavelength calibration can be done either by using separate lamp exposures or by taking several short lamp exposures concurrently with a science exposure. For time-tag data, the latter mode ("tagflash") will allow better correction of potential drift of the spectrum on the detector. One-dimensional spectra will be extracted and saved in a FITS binary table. Separate columns will be used for the flux-calibrated spectrum, error estimate, and the associated wavelengths. CALCOS is written in Python, with some functions in C. It is similar in style to other HST pipeline code in that it uses an association table to specify which files to be included, and the calibration steps to be performed and the reference files to use are specified by header keywords. Currently, in conjunction with the Instrument Definition Team (led by J. Green), the ground-based reference files are being refined, delivered, and tested with the pipeline.

  13. Energy calibration via correlation

    Science.gov (United States)

    Maier, Daniel; Limousin, Olivier

    2016-03-01

    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be less than ~ 0.1 keV. Energy calibration via correlation can be applied to any kind of calibration spectra and shows a robust behavior at low counting statistics. It enables a fast and accurate calibration that can be used to monitor the spectroscopic properties of a detector system in near realtime.

  14. Photogrammetric camera calibration

    Science.gov (United States)

    Tayman, W.P.; Ziemann, H.

    1984-01-01

    Section 2 (Calibration) of the document "Recommended Procedures for Calibrating Photogrammetric Cameras and Related Optical Tests" from the International Archives of Photogrammetry, Vol. XIII, Part 4, is reviewed in the light of recent practical work, and suggestions for changes are made. These suggestions are intended as a basis for a further discussion. ?? 1984.

  15. Sandia WIPP calibration traceability

    Energy Technology Data Exchange (ETDEWEB)

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  16. The GERDA calibration system

    International Nuclear Information System (INIS)

    A system with three identical custom made units is used for the energy calibration of the GERDA Ge diodes. To perform a calibration the 228Th sources are lowered from the parking positions at the top of the cryostat. Their positions are measured by two independent modules. One, the incremental encoder, counts the holes in the perforated steel band holding the sources, the other measures the drive shaft's angular position even if not powered. The system can be controlled remotely by a Labview program. The calibration data is analyzed by an iterative calibration algorithm determining the calibration functions for different energy reconstruction algorithms and the resolution of several peaks in the 228Th spectrum is determined. A Monte Carlo simulation using the GERDA simulation software MAGE has been performed to determine the background induced by the sources in the parking positions.

  17. Sandia WIPP calibration traceability

    International Nuclear Information System (INIS)

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities

  18. Remote calibration of ionization chambers for radioactivity calibration

    International Nuclear Information System (INIS)

    A new calibration technique, referred to as e-trace, has been developed by the National Institute of Advanced Industrial Science and Technology (AIST). The e-trace technique enables rapid remote calibration of measurement equipment and requires minimal resources. We calibrated radioisotope calibrators of the Japan Radioisotope Association (JRIA) and the Nishina Memorial Cyclotron Center (NMCC) remotely and confirmed that remote calibration provided results that are consistent with the results obtained by existing methods within the limits of uncertainty. Accordingly, e-trace has been approved as the standard calibration method at AIST. We intend to apply remote calibration to radioisotope calibrators in hospitals and isotope facilities. (author)

  19. Calibrations of photomultiplier tubes

    International Nuclear Information System (INIS)

    The experimental methods for calibration photomultiplier tubes used in the multichannel fast-pulse-detection system of Thomson scattering measurements for nuclear fusion devices is reported. The most important parameters of the photomultiplier tubes to be calibrated include: linearity of output electric signals to input light signals, response time of pulsed light, spectral response, absolute responsibility, and sensitivity as a function of the chain voltage. The calibrations of all these parameters are carried out by using EMI 9558 B and RCA 7265 photomultiplier tubes respectively. The experimental methods presented in the paper are common to those quantitative measurements that require phomultiplier tubes as detectors

  20. Equipment for dosemeter calibration

    International Nuclear Information System (INIS)

    The device is used for precise calibration of dosimetric instrumentation, such as used at nuclear facilities. The high precision of the calibration procedure is primarily due to the fact that one single and steady radiation source is used. The accurate alignment of the source and the absence of shielding materials in the beam axis make for high homogeneity of the beam and reproducibility of the measurement; this is also contributed to by the horizontal displacement of the optical bench, which ensures a constant temperature field and the possibility of adjusting the radiation source at a sufficient distance from the instrument to be calibrated. (Z.S.). 3 figs

  1. Lidar Calibration Centre

    Science.gov (United States)

    Pappalardo, Gelsomina; Freudenthaler, Volker; Nicolae, Doina; Mona, Lucia; Belegante, Livio; D'Amico, Giuseppe

    2016-06-01

    This paper presents the newly established Lidar Calibration Centre, a distributed infrastructure in Europe, whose goal is to offer services for complete characterization and calibration of lidars and ceilometers. Mobile reference lidars, laboratories for testing and characterization of optics and electronics, facilities for inspection and debugging of instruments, as well as for training in good practices are open to users from the scientific community, operational services and private sector. The Lidar Calibration Centre offers support for trans-national access through the EC HORIZON2020 project ACTRIS-2.

  2. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  3. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina; Webb, Craig

    2016-05-02

    This presentation provides a high-level overview of the progress on the Broadband Outdoor Radiometer Calibrations for all shortwave and longwave radiometers that are deployed by the Atmospheric Radiation Measurement program.

  4. Air Data Calibration Facility

    Data.gov (United States)

    Federal Laboratory Consortium — This facility is for low altitude subsonic altimeter system calibrations of air vehicles. Mission is a direct support of the AFFTC mission. Postflight data merge is...

  5. Approximation Behooves Calibration

    DEFF Research Database (Denmark)

    da Silva Ribeiro, André Manuel; Poulsen, Rolf

    2013-01-01

    Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009.......Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009....

  6. Scanner calibration revisited

    Directory of Open Access Journals (Sweden)

    Pozhitkov Alexander E

    2010-07-01

    Full Text Available Abstract Background Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2. reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al. Methods Signal intensities were recorded on three different microarray scanners at various photomultiplier gain levels using the same calibration slide from Full Moon BioSystems. Data analysis was conducted on raw signal intensities without normalization or transformation of any kind. Weighted least-squares method was used to fit the data. Results We found that initial analysis performed by Shi et al. did not take into account autofluorescence of the Full Moon BioSystems slide, which led to a grossly distorted microarray scanner response. Our analysis revealed that a power-law function, which is explicitly accounting for the slide autofluorescence, perfectly described a relationship between signal intensities and fluorophore quantities. Conclusions Microarray scanners respond in a much less distorted fashion than was reported by Shi et al. Full Moon BioSystems calibration slides are inadequate for performing calibration. We recommend against using these slides.

  7. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  8. Energy calibration via correlation

    CERN Document Server

    Maier, Daniel

    2015-01-01

    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241 Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be le...

  9. Calibration of sound velocimeter in pure water

    Science.gov (United States)

    Li, Zhiwei; Zhang, Baofeng; Li, Tao; Zhu, Junchao; Xie, Ziming

    2016-01-01

    Accurate measurement of sound speed is important to calibrate a sound velocity profiler which provides real-time sound velocity to the sonar equipment in oceanographic survey. The sound velocity profiler calculates the sound speed by measuring the time-of-flight of a 1 MHz single acoustic pulse to travel over about 300 mm path. A standard sound velocimeter instrument was invited to calibrate the sound velocity profiler in pure water at temperatures of 278,283, 288, 293, 298, 303 and 308K in a thermostatic vessel at one atmosphere. The sound velocity profiler was deployed in the thermostatic vessel alongside the standard sound velocimeter instrument and two platinum resistance thermometers (PRT) which were calibrated to 0.002k by comparison with a standard PRT. Time of flight circuit board was used to measure the time-of-flight to 22 picosecond precision. The sound speed which was measured by the sound velocity profiler was compared to the standard sound speed calculated by UNESCO to give the laboratory calibration coefficients and was demonstrated agreement with CTD-derived sound speed using Del Grosso's seawater equation after removing a bias.

  10. HAWC Timing Calibration

    CERN Document Server

    Huentemeyer, Petra; Dingus, Brenda

    2009-01-01

    The High-Altitude Water Cherenkov (HAWC) Experiment is a second-generation highsensitivity gamma-ray and cosmic-ray detector that builds on the experience and technology of the Milagro observatory. Like Milagro, HAWC utilizes the water Cherenkov technique to measure extensive air showers. Instead of a pond filled with water (as in Milagro) an array of closely packed water tanks is used. The event direction will be reconstructed using the times when the PMTs in each tank are triggered. Therefore, the timing calibration will be crucial for reaching an angular resolution as low as 0.25 degrees.We propose to use a laser calibration system, patterned after the calibration system in Milagro. Like Milagro, the HAWC optical calibration system will use ~1 ns laser light pulses. Unlike Milagro, the PMTs are optically isolated and require their own optical fiber calibration. For HAWC the laser light pulses will be directed through a series of optical fan-outs and fibers to illuminate the PMTs in approximately one half o...

  11. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  12. GTC Photometric Calibration

    Science.gov (United States)

    di Cesare, M. A.; Hammersley, P. L.; Rodriguez Espinosa, J. M.

    2006-06-01

    We are currently developing the calibration programme for GTC using techniques similar to the ones use for the space telescope calibration (Hammersley et al. 1998, A&AS, 128, 207; Cohen et al. 1999, AJ, 117, 1864). We are planning to produce a catalogue with calibration stars which are suitable for a 10-m telescope. These sources will be not variable, non binary and do not have infrared excesses if they are to be used in the infrared. The GTC science instruments require photometric calibration between 0.35 and 2.5 microns. The instruments are: OSIRIS (Optical System for Imaging low Resolution Integrated Spectroscopy), ELMER and EMIR (Espectrógrafo Multiobjeto Infrarrojo) and the Acquisition and Guiding boxes (Di Césare, Hammersley, & Rodriguez Espinosa 2005, RevMexAA Ser. Conf., 24, 231). The catalogue will consist of 30 star fields distributed in all of North Hemisphere. We will use fields containing sources over the range 12 to 22 magnitude, and spanning a wide range of spectral types (A to M) for the visible and near infrared. In the poster we will show the method used for selecting these fields and we will present the analysis of the data on the first calibration fields observed.

  13. Calibrating nacelle lidars

    DEFF Research Database (Denmark)

    Courtney, Michael

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report...... accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a...... inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work....

  14. TARGETLESS CAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    L. Barazzetti

    2012-09-01

    Full Text Available In photogrammetry a camera is considered calibrated if its interior orientation parameters are known. These encompass the principal distance, the principal point position and some Additional Parameters used to model possible systematic errors. The current state of the art for automated camera calibration relies on the use of coded targets to accurately determine the image correspondences. This paper presents a new methodology for the efficient and rigorous photogrammetric calibration of digital cameras which does not require any longer the use of targets. A set of images depicting a scene with a good texture are sufficient for the extraction of natural corresponding image points. These are automatically matched with feature-based approaches and robust estimation techniques. The successive photogrammetric bundle adjustment retrieves the unknown camera parameters and their theoretical accuracies. Examples, considerations and comparisons with real data and different case studies are illustrated to show the potentialities of the proposed methodology.

  15. The MINOS calibration detector

    International Nuclear Information System (INIS)

    This paper describes the MINOS calibration detector (CalDet) and the procedure used to calibrate it. The CalDet, a scaled-down but functionally equivalent model of the MINOS Far and Near detectors, was exposed to test beams in the CERN PS East Area during 2001-2003 to establish the response of the MINOS calorimeters to hadrons, electrons and muons in the range 0.2-10GeV/c. The CalDet measurements are used to fix the energy scale and constrain Monte Carlo simulations of MINOS

  16. Individual dosimetry and calibration

    International Nuclear Information System (INIS)

    In 1995 both the Individual Dosimetry and Calibration Sections worked under the condition of a status quo and concentrated fully on the routine part of their work. Nevertheless, the machine for printing the bar code which will be glued onto the film holder and hence identify the people when entering into high radiation areas was put into operation and most of the holders were equipped with the new identification. As far as the Calibration Section is concerned the project of the new source control system that is realized by the Technical Support Section was somewhat accelerated

  17. Calibration of farmer dosemeters

    International Nuclear Information System (INIS)

    The Farmer Dosemeters of Atomic Energy Medical Centre (AEMC) Jamshoro were calibrated in the Secondary Standard Dosimetry Laboratory (SSDL) at PINSTECH, using the NPL Secondary Standard Therapy level X-ray exposure meter. The results are presented in this report. (authors)

  18. Calibration Of Oxygen Monitors

    Science.gov (United States)

    Zalenski, M. A.; Rowe, E. L.; Mcphee, J. R.

    1988-01-01

    Readings corrected for temperature, pressure, and humidity of air. Program for handheld computer developed to ensure accuracy of oxygen monitors in National Transonic Facility, where liquid nitrogen stored. Calibration values, determined daily, based on entries of data on barometric pressure, temperature, and relative humidity. Output provided directly in millivolts.

  19. Commodity-Free Calibration

    Science.gov (United States)

    2008-01-01

    Commodity-free calibration is a reaction rate calibration technique that does not require the addition of any commodities. This technique is a specific form of the reaction rate technique, where all of the necessary reactants, other than the sample being analyzed, are either inherent in the analyzing system or specifically added or provided to the system for a reason other than calibration. After introduction, the component of interest is exposed to other reactants or flow paths already present in the system. The instrument detector records one of the following to determine the rate of reaction: the increase in the response of the reaction product, a decrease in the signal of the analyte response, or a decrease in the signal from the inherent reactant. With this data, the initial concentration of the analyte is calculated. This type of system can analyze and calibrate simultaneously, reduce the risk of false positives and exposure to toxic vapors, and improve accuracy. Moreover, having an excess of the reactant already present in the system eliminates the need to add commodities, which further reduces cost, logistic problems, and potential contamination. Also, the calculations involved can be simplified by comparison to those of the reaction rate technique. We conducted tests with hypergols as an initial investigation into the feasiblility of the technique.

  20. Calibration bench of flowmeters

    International Nuclear Information System (INIS)

    This equipment is devoted to the comparison of signals from two turbines installed in the Cabri experimental loop. The signal is compared to the standard turbine. The characteristics and the performance of the calibration bench are presented. (A.L.B.)

  1. Measurement System & Calibration report

    DEFF Research Database (Denmark)

    Vesth, Allan; Kock, Carsten Weber

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  2. Calibration with Absolute Shrinkage

    DEFF Research Database (Denmark)

    Øjelund, Henrik; Madsen, Henrik; Thyregod, Poul

    2001-01-01

    In this paper, penalized regression using the L-1 norm on the estimated parameters is proposed for chemometric je calibration. The algorithm is of the lasso type, introduced by Tibshirani in 1996 as a linear regression method with bound on the absolute length of the parameters, but a modification...

  3. Calibration issues for MUSE

    Science.gov (United States)

    Kelz, Andreas; Roth, Martin; Bauer, Svend; Gerssen, Joris; Hahn, Thomas; Weilbacher, Peter; Laux, Uwe; Loupias, Magali; Kosmalski, Johan; McDermid, Richard; Bacon, Roland

    2008-07-01

    The Multi-Unit Spectroscopic Explorer (MUSE) is an integral-field spectrograph for the VLT for the next decade. Using an innovative field-splitting and slicing design, combined with an assembly of 24 spectrographs, MUSE will provide some 90,000 spectra in one exposure, which cover a simultaneous spectral range from 465 to 930nm. The design and manufacture of the Calibration Unit, the alignment tests of the Spectrograph and Detector sub-systems, and the development of the Data Reduction Software for MUSE are work-packages under the responsibility of the AIP, who is a partner in a European-wide consortium of 6 institutes and ESO, that is led by the Centre de Recherche Astronomique de Lyon. MUSE will be operated and therefore has to be calibrated in a variety of modes, which include seeing-limited and AO-assisted operations, providing a wide and narrow-field-of-view. MUSE aims to obtain unprecedented ultra-deep 3D-spectroscopic exposures, involving integration times of the order of 80 hours at the VLT. To achieve the corresponding science goals, instrumental stability, accurate calibration and adequate data reduction tools are needed. The paper describes the status at PDR of the AIP related work-packages, in particular with respect to the spatial, spectral, image quality, and geometrical calibration and related data reduction aspects.

  4. Entropic calibration revisited

    Energy Technology Data Exchange (ETDEWEB)

    Brody, Dorje C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)]. E-mail: d.brody@imperial.ac.uk; Buckley, Ian R.C. [Centre for Quantitative Finance, Imperial College, London SW7 2AZ (United Kingdom); Constantinou, Irene C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom); Meister, Bernhard K. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)

    2005-04-11

    The entropic calibration of the risk-neutral density function is effective in recovering the strike dependence of options, but encounters difficulties in determining the relevant greeks. By use of put-call reversal we apply the entropic method to the time reversed economy, which allows us to obtain the spot price dependence of options and the relevant greeks.

  5. Physiotherapy ultrasound calibrations

    International Nuclear Information System (INIS)

    Calibration of physiotherapy ultrasound equipment has long been a problem. Numerous surveys around the world over the past 20 years have all found that only a low percentage of the units tested had an output within 30% of that indicatd. In New Zealand, a survey carried out by the NRL in 1985 found that only 24% had an output, at the maximum setting, within + or - 20% of that indicated. The present performance Standard for new equipment (NZS 3200.2.5:1992) requires that the measured output should not deviate from that indicated by more than + or - 30 %. This may be tightened to + or - 20% in the next few years. Any calibration is only as good as the calibration equipment. Some force balances can be tested with small weights to simulate the force exerted by an ultrasound beam, but with others this is not possible. For such balances, testing may only be feasible with a calibrated source which could be used like a transfer standard. (author). 4 refs., 3 figs

  6. NVLAP calibration laboratory program

    Energy Technology Data Exchange (ETDEWEB)

    Cigler, J.L.

    1993-12-31

    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  7. LOFAR Facet Calibration

    Science.gov (United States)

    van Weeren, R. J.; Williams, W. L.; Hardcastle, M. J.; Shimwell, T. W.; Rafferty, D. A.; Sabater, J.; Heald, G.; Sridhar, S. S.; Dijkema, T. J.; Brunetti, G.; Brüggen, M.; Andrade-Santos, F.; Ogrean, G. A.; Röttgering, H. J. A.; Dawson, W. A.; Forman, W. R.; de Gasperin, F.; Jones, C.; Miley, G. K.; Rudnick, L.; Sarazin, C. L.; Bonafede, A.; Best, P. N.; Bîrzan, L.; Cassano, R.; Chyży, K. T.; Croston, J. H.; Ensslin, T.; Ferrari, C.; Hoeft, M.; Horellou, C.; Jarvis, M. J.; Kraft, R. P.; Mevius, M.; Intema, H. T.; Murray, S. S.; Orrú, E.; Pizzo, R.; Simionescu, A.; Stroe, A.; van der Tol, S.; White, G. J.

    2016-03-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction-dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction-dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at ∼ 5\\prime\\prime resolution, meeting the specifications of the LOFAR Tier-1 northern survey.

  8. Radiation monitor calibration technique

    International Nuclear Information System (INIS)

    Reference radiations in the Secondary Standard Dosimetry Laboratory, OAEP have been improved and modified by employing lead attenuators. To identify low-level exposure rate, shadow-cone method has been applied. The secondary standard dosemeter has been used periodically to check the constancy of reference radiations to assure the calibration of dosemeters and dose-ratemeters used for radiation protection

  9. LOFAR facet calibration

    CERN Document Server

    van Weeren, R J; Hardcastle, M J; Shimwell, T W; Rafferty, D A; Sabater, J; Heald, G; Sridhar, S S; Dijkema, T J; Brunetti, G; Brüggen, M; Andrade-Santos, F; Ogrean, G A; Röttgering, H J A; Dawson, W A; Forman, W R; de Gasperin, F; Jones, C; Miley, G K; Rudnick, L; Sarazin, C L; Bonafede, A; Best, P N; Bîrzan, L; Cassano, R; Chyży, K T; Croston, J H; Ensslin, T; Ferrari, C; Hoeft, M; Horellou, C; Jarvis, M J; Kraft, R P; Mevius, M; Intema, H T; Murray, S S; Orrú, E; Pizzo, R; Simionescu, A; Stroe, A; van der Tol, S; White, G J

    2016-01-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at $\\sim$ 5arcsec resolu...

  10. Pseudo Linear Gyro Calibration

    Science.gov (United States)

    Harman, Richard; Bar-Itzhack, Itzhack Y.

    2003-01-01

    Previous high fidelity onboard attitude algorithms estimated only the spacecraft attitude and gyro bias. The desire to promote spacecraft and ground autonomy and improvements in onboard computing power has spurred development of more sophisticated calibration algorithms. Namely, there is a desire to provide for sensor calibration through calibration parameter estimation onboard the spacecraft as well as autonomous estimation on the ground. Gyro calibration is a particularly challenging area of research. There are a variety of gyro devices available for any prospective mission ranging from inexpensive low fidelity gyros with potentially unstable scale factors to much more expensive extremely stable high fidelity units. Much research has been devoted to designing dedicated estimators such as particular Extended Kalman Filter (EKF) algorithms or Square Root Information Filters. This paper builds upon previous attitude, rate, and specialized gyro parameter estimation work performed with Pseudo Linear Kalman Filter (PSELIKA). The PSELIKA advantage is the use of the standard linear Kalman Filter algorithm. A PSELIKA algorithm for an orthogonal gyro set which includes estimates of attitude, rate, gyro misalignments, gyro scale factors, and gyro bias is developed and tested using simulated and flight data. The measurements PSELIKA uses include gyro and quaternion tracker data.

  11. Pleiades Absolute Calibration : Inflight Calibration Sites and Methodology

    Science.gov (United States)

    Lachérade, S.; Fourest, S.; Gamet, P.; Lebègue, L.

    2012-07-01

    In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station) and Oceans (Calibration over molecular scattering) or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  12. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini

    2007-03-31

    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  13. Calibrated Properties Model

    International Nuclear Information System (INIS)

    The purpose of this Model Report is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Office of Repository Development (ORD). The UZ contains the unsaturated rock layers overlying the repository and host unit, which constitute a natural barrier to flow, and the unsaturated rock layers below the repository which constitute a natural barrier to flow and transport. This work followed, and was planned in, ''Technical Work Plan (TWP) for: Performance Assessment Unsaturated Zone'' (BSC 2002 [160819], Section 1.10.8 [under Work Package (WP) AUZM06, Climate Infiltration and Flow], and Section I-1-1 [in Attachment I, Model Validation Plans]). In Section 4.2, four acceptance criteria (ACs) are identified for acceptance of this Model Report; only one of these (Section 4.2.1.3.6.3, AC 3) was identified in the TWP (BSC 2002 [160819], Table 3-1). These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, and drift-scale and mountain-scale coupled-process models from the UZ Flow, Transport and Coupled Processes Department in the Natural Systems Subproject of the Performance Assessment (PA) Project. The Calibrated Properties Model output will also be used by the Engineered Barrier System Department in the Engineering Systems Subproject. The Calibrated Properties Model provides input through the UZ Model and other process models of natural and engineered systems to the Total System Performance Assessment (TSPA) models, in accord with the PA Strategy and Scope in the PA Project of the Bechtel SAIC Company, LLC (BSC). The UZ process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions. UZ flow is a TSPA model component

  14. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten;

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve the...... statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited for...

  15. Mercury Calibration System

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Joseph Rovani; Mark Sanderson; Ryan Boysen; William Schuster

    2009-03-11

    U.S. Environmental Protection Agency (EPA) Performance Specification 12 in the Clean Air Mercury Rule (CAMR) states that a mercury CEM must be calibrated with National Institute for Standards and Technology (NIST)-traceable standards. In early 2009, a NIST traceable standard for elemental mercury CEM calibration still does not exist. Despite the vacature of CAMR by a Federal appeals court in early 2008, a NIST traceable standard is still needed for whatever regulation is implemented in the future. Thermo Fisher is a major vendor providing complete integrated mercury continuous emissions monitoring (CEM) systems to the industry. WRI is participating with EPA, EPRI, NIST, and Thermo Fisher towards the development of the criteria that will be used in the traceability protocols to be issued by EPA. An initial draft of an elemental mercury calibration traceability protocol was distributed for comment to the participating research groups and vendors on a limited basis in early May 2007. In August 2007, EPA issued an interim traceability protocol for elemental mercury calibrators. Various working drafts of the new interim traceability protocols were distributed in late 2008 and early 2009 to participants in the Mercury Standards Working Committee project. The protocols include sections on qualification and certification. The qualification section describes in general terms tests that must be conducted by the calibrator vendors to demonstrate that their calibration equipment meets the minimum requirements to be established by EPA for use in CAMR monitoring. Variables to be examined include linearity, ambient temperature, back pressure, ambient pressure, line voltage, and effects of shipping. None of the procedures were described in detail in the draft interim documents; however they describe what EPA would like to eventually develop. WRI is providing the data and results to EPA for use in developing revised experimental procedures and realistic acceptance criteria based on

  16. Streak camera time calibration procedures

    Science.gov (United States)

    Long, J.; Jackson, I.

    1978-01-01

    Time calibration procedures for streak cameras utilizing a modulated laser beam are described. The time calibration determines a writing rate accuracy of 0.15% with a rotating mirror camera and 0.3% with an image converter camera.

  17. The Calibration Reference Data System

    Science.gov (United States)

    Greenfield, P.; Miller, T.

    2016-07-01

    We describe a software architecture and implementation for using rules to determine which calibration files are appropriate for calibrating a given observation. This new system, the Calibration Reference Data System (CRDS), replaces what had been previously used for the Hubble Space Telescope (HST) calibration pipelines, the Calibration Database System (CDBS). CRDS will be used for the James Webb Space Telescope (JWST) calibration pipelines, and is currently being used for HST calibration pipelines. CRDS can be easily generalized for use in similar applications that need a rules-based system for selecting the appropriate item for a given dataset; we give some examples of such generalizations that will likely be used for JWST. The core functionality of the Calibration Reference Data System is available under an Open Source license. CRDS is briefly contrasted with a sampling of other similar systems used at other observatories.

  18. Optical tweezers absolute calibration

    CERN Document Server

    Dutra, R S; Neto, P A Maia; Nussenzveig, H M

    2014-01-01

    Optical tweezers are highly versatile laser traps for neutral microparticles, with fundamental applications in physics and in single molecule cell biology. Force measurements are performed by converting the stiffness response to displacement of trapped transparent microspheres, employed as force transducers. Usually, calibration is indirect, by comparison with fluid drag forces. This can lead to discrepancies by sizable factors. Progress achieved in a program aiming at absolute calibration, conducted over the past fifteen years, is briefly reviewed. Here we overcome its last major obstacle, a theoretical overestimation of the peak stiffness, within the most employed range for applications, and we perform experimental validation. The discrepancy is traced to the effect of primary aberrations of the optical system, which are now included in the theory. All required experimental parameters are readily accessible. Astigmatism, the dominant effect, is measured by analyzing reflected images of the focused laser spo...

  19. Astrid-2 SSC ASUMagnetic Calibration

    DEFF Research Database (Denmark)

    Primdahl, Fritz

    1997-01-01

    Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory.......Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory....

  20. Optical Calibration of SNO+

    Science.gov (United States)

    Maneira, J.; Peeters, S.; Sinclair, J.

    2015-04-01

    SNO is being upgraded to SNO+, which has as its main goal the search for neutrinoless double-beta decay. The upgrade is defined by filling with a novel scintillator mixture containing 130Te. With a lower energy threshold than SNO, SNO+ will be sensitive to other exciting new physics. Here we are describing new optical calibration system that meets new, more stringent radiopurity requirements has been developed.

  1. Camera Calibration Using Silhouettes

    OpenAIRE

    Boyer, Edmond

    2005-01-01

    This report addresses the problem of estimating camera parameters from images where object silhouettes only are known. Several modeling applications make use of silhouettes, and while calibration methods are well known when considering points or lines matched along image sequences, the problem appears to be more difficult when considering silhouettes. However, such primitives encode also information on camera parameters by the fact that their associated viewing cones should present a common i...

  2. Program Calibrates Strain Gauges

    Science.gov (United States)

    Okazaki, Gary D.

    1991-01-01

    Program dramatically reduces personnel and time requirements for acceptance tests of hardware. Data-acquisition system reads output from Wheatstone full-bridge strain-gauge circuit and calculates strain by use of shunt calibration technique. Program nearly instantaneously tabulates and plots strain data against load-cell outputs. Modified to acquire strain data for other specimens wherever full-bridge strain-gauge circuits used. Written in HP BASIC.

  3. Calibration specimens for microscopy

    Czech Academy of Sciences Publication Activity Database

    Kolařík, Vladimír; Matějka, Milan; Matějka, František; Krátký, Stanislav; Urbánek, Michal; Horáček, Miroslav; Král, Stanislav; Bok, Jan

    Ostrava: TANGER Ltd, 2012, s. 713-716. ISBN 978-80-87294-32-1. [NANOCON 2012. International Conference /4./. Brno (CZ), 23.10.2012-25.10.2012] R&D Projects: GA MŠk ED0017/01/01; GA TA ČR TE01020233; GA MPO FR-TI1/576 Institutional support: RVO:68081731 Keywords : E-beam technology * calibration specimen * scanning electron microscopy Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering

  4. Calibrating bacterial evolution

    OpenAIRE

    Ochman, Howard; Elwyn, Susannah; Moran, Nancy A

    1999-01-01

    Attempts to calibrate bacterial evolution have relied on the assumption that rates of molecular sequence divergence in bacteria are similar to those of higher eukaryotes, or to those of the few bacterial taxa for which ancestors can be reliably dated from ecological or geological evidence. Despite similarities in the substitution rates estimated for some lineages, comparisons of the relative rates of evolution at different classes of nucleotide sites indicate no basis for their universal appl...

  5. Mesoscale hybrid calibration artifact

    Science.gov (United States)

    Tran, Hy D.; Claudet, Andre A.; Oliver, Andrew D.

    2010-09-07

    A mesoscale calibration artifact, also called a hybrid artifact, suitable for hybrid dimensional measurement and the method for make the artifact. The hybrid artifact has structural characteristics that make it suitable for dimensional measurement in both vision-based systems and touch-probe-based systems. The hybrid artifact employs the intersection of bulk-micromachined planes to fabricate edges that are sharp to the nanometer level and intersecting planes with crystal-lattice-defined angles.

  6. Dosimetry and Calibration Section

    International Nuclear Information System (INIS)

    The two tasks of the Dosimetry and Calibration Section at CERN are the Individual Dosimetry Service which assures the personal monitoring of about 5000 persons potentially exposed to ionizing radiation at CERN, and the Calibration Laboratory which verifies all the instruments and monitors. This equipment is used by the sections of the RP Group for assuring radiation protection around CERN's accelerators, and by the Environmental Section of TISTE. In addition, nearly 250 electronic and 300 quartz fibre dosimeters, employed in operational dosimetry, are calibrated at least once a year. The Individual Dosimetry Service uses an extended database (INDOS) which contains information about all the individual doses ever received at CERN. For most of 1997 it was operated without the support of a database administrator as the technician who had assured this work retired. The Software Support Section of TIS-TE took over the technical responsibility of the database, but in view of the many other tasks of this Section and the lack of personnel, only a few interventions for solving immediate problems were possible

  7. Calibrated Properties Model

    International Nuclear Information System (INIS)

    The purpose of this model report is to document the calibrated properties model that provides calibrated property sets for unsaturated zone (UZ) flow and transport process models (UZ models). The calibration of the property sets is performed through inverse modeling. This work followed, and was planned in, ''Technical Work Plan (TWP) for: Unsaturated Zone Flow Analysis and Model Report Integration'' (BSC 2004 [DIRS 169654], Sections 1.2.6 and 2.1.1.6). Direct inputs to this model report were derived from the following upstream analysis and model reports: ''Analysis of Hydrologic Properties Data'' (BSC 2004 [DIRS 170038]); ''Development of Numerical Grids for UZ Flow and Transport Modeling'' (BSC 2004 [DIRS 169855]); ''Simulation of Net Infiltration for Present-Day and Potential Future Climates'' (BSC 2004 [DIRS 170007]); ''Geologic Framework Model'' (GFM2000) (BSC 2004 [DIRS 170029]). Additionally, this model report incorporates errata of the previous version and closure of the Key Technical Issue agreement TSPAI 3.26 (Section 6.2.2 and Appendix B), and it is revised for improved transparency

  8. Del Piero

    OpenAIRE

    Umbaca, Enzo

    2015-01-01

    Rifacendosi al “Pinturicchio”, epiteto con cui Gianni Agnelli consacrava il calciatore Alessandro Del Piero un “artista del calcio”, Enzo Umbaca scrisse al popolare personaggio invitandolo a collaborare ad una performance durante la quale Pinturicchio-Del Piero avrebbe affrescato il muro di una galleria torinese calciando un pallone macchiato di grafite contro un muro sul quale è appesa la copia di un opera del Pinturicchio.

  9. Calibration effects on orbit determination

    Science.gov (United States)

    Madrid, G. A.; Winn, F. B.; Zielenbach, J. W.; Yip, K. B.

    1974-01-01

    The effects of charged particle and tropospheric calibrations on the orbit determination (OD) process are analyzed. The calibration process consisted of correcting the Doppler observables for the media effects. Calibrated and uncalibrated Doppler data sets were used to obtain OD results for past missions as well as Mariner Mars 1971. Comparisons of these Doppler reductions show the significance of the calibrations. For the MM'71 mission, the media calibrations proved themselves effective in diminishing the overall B-plane error and reducing the Doppler residual signatures.

  10. Mammography calibration: Factor or fit?

    International Nuclear Information System (INIS)

    Dose measurements in mammography x-ray have become more important and a basic path in quality assurance programmes. It is recognized by the international guidelines that it is necessary to have calibration services offered for mammography beams in order to help the improvement of the clinical diagnosis. Major efforts have been made by several laboratories in order to establish an appropriate and traceable calibration infrastructure and to provide the basis for a quality control programme in mammography. The indication of a dosimeter, whose reference point is positioned at the point of test, is compared with the conventional true value of the quantity to be measured. The calibration coefficient is then the ratio of the conventional true value to the indicated. The Primary Standard Dosimetry Laboratory - PSDL or the Secondary Standard Dosimetry Laboratory - SSDL provides the calibration coefficient of the dosimeters in reference to the Half Value Layers - HVL implemented in their laboratories. The dosimetry calibration data is enough when the user has the same system as the laboratory where the ionization chamber has been calibrated. However, there are other calibration systems that have different calibration qualities implemented using different combinations of anode and filter and, therefore, there is no direct relation with the calibration coefficient. How to deal with this? There are two different ways to obtain calibration coefficients when the user's implemented qualities are different from the calibration laboratory's qualities. The first is the interpolation of each calibration coefficient stated in the certificate. The second is the fit of all calibration coefficients, separately for non-attenuated and attenuated beam qualities, to obtain a function by which the calibration coefficients can be determined at each beam quality. The second one includes the statistical fluctuation. The dosimetry calibration data must fit an analytical form, as for example a

  11. A variable acceleration calibration system

    Science.gov (United States)

    Johnson, Thomas H.

    2011-12-01

    A variable acceleration calibration system that applies loads using gravitational and centripetal acceleration serves as an alternative, efficient and cost effective method for calibrating internal wind tunnel force balances. Two proof-of-concept variable acceleration calibration systems are designed, fabricated and tested. The NASA UT-36 force balance served as the test balance for the calibration experiments. The variable acceleration calibration systems are shown to be capable of performing three component calibration experiments with an approximate applied load error on the order of 1% of the full scale calibration loads. Sources of error are indentified using experimental design methods and a propagation of uncertainty analysis. Three types of uncertainty are indentified for the systems and are attributed to prediction error, calibration error and pure error. Angular velocity uncertainty is shown to be the largest indentified source of prediction error. The calibration uncertainties using a production variable acceleration based system are shown to be potentially equivalent to current methods. The production quality system can be realized using lighter materials and a more precise instrumentation. Further research is needed to account for balance deflection, forcing effects due to vibration, and large tare loads. A gyroscope measurement technique is shown to be capable of resolving the balance deflection angle calculation. Long term research objectives include a demonstration of a six degree of freedom calibration, and a large capacity balance calibration.

  12. Automatic force balance calibration system

    Science.gov (United States)

    Ferris, Alice T.

    1995-05-01

    A system for automatically calibrating force balances is provided. The invention uses a reference balance aligned with the balance being calibrated to provide superior accuracy while minimizing the time required to complete the calibration. The reference balance and the test balance are rigidly attached together with closely aligned moment centers. Loads placed on the system equally effect each balance, and the differences in the readings of the two balances can be used to generate the calibration matrix for the test balance. Since the accuracy of the test calibration is determined by the accuracy of the reference balance and current technology allows for reference balances to be calibrated to within +/-0.05% the entire system has an accuracy of +/-0.2%. The entire apparatus is relatively small and can be mounted on a movable base for easy transport between test locations. The system can also accept a wide variety of reference balances, thus allowing calibration under diverse load and size requirements.

  13. Structured light camera calibration

    Science.gov (United States)

    Garbat, P.; Skarbek, W.; Tomaszewski, M.

    2013-03-01

    Structured light camera which is being designed with the joined effort of Institute of Radioelectronics and Institute of Optoelectronics (both being large units of the Warsaw University of Technology within the Faculty of Electronics and Information Technology) combines various hardware and software contemporary technologies. In hardware it is integration of a high speed stripe projector and a stripe camera together with a standard high definition video camera. In software it is supported by sophisticated calibration techniques which enable development of advanced application such as real time 3D viewer of moving objects with the free viewpoint or 3D modeller for still objects.

  14. Smart Calibration of Excavators

    DEFF Research Database (Denmark)

    Bro, Marie; Døring, Kasper; Ellekilde, Lars-Peter; Hansen, Mikael Sonne; Markvorsen, Steen; Spence, David; Stolpe, Mathias; Sølvason, Dorthe

    2005-01-01

    Excavators dig holes. But where is the bucket? The purpose of this report is to treat four different problems concerning calibrations of position indicators for excavators in operation at concrete construction sites. All four problems are related to the question of how to determine the precise...... geographic and/or site-relative position of a given excavator and its bucket. However, our presentations and solutions to the problems can, nevertheless, be read and studied in any order and independently of each other. This also implies and induces a gentle warning to the reader: The {\\em{notation}} need...

  15. Use of Radiometrically Calibrated Flat-Plate Calibrators in Calibration of Radiation Thermometers

    Science.gov (United States)

    Cárdenas-García, D.; Méndez-Lango, E.

    2015-08-01

    Most commonly used, low-temperature, infrared thermometers have large fields of view sizes that make them difficult to be calibrated with narrow aperture blackbodies. Flat-plate calibrators with large emitting surfaces have been proposed for calibrating these infrared thermometers. Because the emissivity of the flat plate is not unity, its radiance temperature is wavelength dependent. For calibration, the wavelength pass band of the device under test should match that of the reference infrared thermometer. If the device under test and reference radiometer have different pass bands, then it is possible to calculate the corresponding correction if the emissivity of the flat plate is known. For example, a correction of at is required when calibrating a infrared thermometer with a "" radiometrically calibrated flat-plate calibrator. A method is described for using a radiometrically calibrated flat-plate calibrator that covers both cases of match and mismatch working wavelength ranges of a reference infrared thermometer and infrared thermometers to be calibrated with the flat-plate calibrator. Also, an application example is included in this paper.

  16. Dosimetry and Calibration Section

    International Nuclear Information System (INIS)

    The Dosimetry and Calibration Section fulfils two tasks within CERN's Radiation Protection Group: the Individual Dosimetry Service monitors more than 5000 persons potentially exposed to ionizing radiation on the CERN sites, and the Calibration Laboratory verifies throughout the year, at regular intervals, over 1000 instruments, monitors, and electronic dosimeters used by RP Group. The establishment of a Quality Assurance System for the Individual Dosimetry Service, a requirement of the new Swiss Ordinance for personal dosimetry, put a considerable workload on the section. Together with an external consultant it was decided to identify and then describe the different 'processes' of the routine work performed in the dosimetry service. The resulting Quality Manual was submitted to the Federal Office for Public Health in Bern in autumn. The CERN Individual Dosimetry Service will eventually be officially endorsed after a successful technical test in March 1999. On the technical side, the introduction of an automatic development machine for gamma films was very successful. It processes the dosimetric films without an operator being present, and its built-in regeneration mechanism keeps the concentration of the processing chemicals at a constant level

  17. NIR prediction of fruit moisture, free acidity and oil content in intact olives

    Directory of Open Access Journals (Sweden)

    2009-06-01

    Full Text Available In this work, the prediction of olive fruit and virgin olive oil quality parameters through the direct measuring of the fruit using near infrared spectrometry (NIRS has been investigated and the effectiveness of a portable spectrometer has been assessed. Models and calibration tests were developed using both the hexane-isopropanol extraction of individual olive fruits, and the Soxhlet extraction of olive paste. The parameters analyzed were the free acidity in olive oil, oil yield from physical extraction, oil content referring to fresh weight, oil content referring to dry matter and fruit moisture. The results indicate a good predictive potential with both methodologies and serve to encourage improvement in the obtained models through the enlargement of the calibrations. Fruit moisture prediction models showed high accuracy.

    En este trabajo se ha investigado la predicción de parámetros de calidad de aceitunas y de aceite de oliva virgen mediante medidas directas en el fruto de espectrometría de infrarrojo cercano (NIRS, evaluándose la utilidad de un espectrómetro portátil. Se han desarrollado respectivamente modelos predictivos y calibraciones utilizando como análisis de referencia tanto la extracción de aceitunas individualmente con hexano-isopropanol, como la extracción de pasta de aceituna mediante Soxhlet. Los parámetros analizados fueron: acidez libre del aceite, rendimiento de la extracción física de aceite, contenido de aceite referido a peso fresco, contenido de aceite referido a materia seca y humedad del fruto. Los resultados indican un buen potencial de predicción mediante ambos métodos y alientan al perfeccionamiento de los modelos obtenidos mediante la ampliación de las calibraciones. Los modelos predictivos de la humedad del fruto mostraron una alta precisión.

  18. The Detector Calibration System for the CUORE cryogenic bolometer array

    CERN Document Server

    Cushman, J S; Davis, C J; Ejzak, L; Lenz, D; Lim, K E; Heeger, K M; Maruyama, R H; Nucciotti, A; Sangiorgio, S; Wise, T

    2016-01-01

    The Cryogenic Underground Observatory for Rare Events (CUORE) is a ton-scale cryogenic experiment designed to search for neutrinoless double-beta decay of $^{130}$Te and other rare events. The CUORE detector consists of 988 TeO$_2$ bolometers operated underground at 10~mK in a dilution refrigerator at the Laboratori Nazionali del Gran Sasso. Candidate events are identified through a precise measurement of their energy. The absolute energy response of the detectors is established by the regular calibration of each individual bolometer using gamma sources. The close-packed configuration of the CUORE bolometer array combined with the extensive shielding surrounding the detectors requires the placement of calibration sources within the array itself. The CUORE Detector Calibration System is designed to insert radioactive sources into and remove them from the cryostat while respecting the stringent heat load, radiopurity, and operational requirements of the experiment. This paper describes the design, commissioning...

  19. Camera Calibration: a USU Implementation

    OpenAIRE

    Ma, Lili; Chen, YangQuan; Moore, Kevin L.

    2003-01-01

    The task of camera calibration is to estimate the intrinsic and extrinsic parameters of a camera model. Though there are some restricted techniques to infer the 3-D information about the scene from uncalibrated cameras, effective camera calibration procedures will open up the possibility of using a wide range of existing algorithms for 3-D reconstruction and recognition. The applications of camera calibration include vision-based metrology, robust visual platooning and visual docking of mobil...

  20. Calibration procedure for zenith plummets

    Directory of Open Access Journals (Sweden)

    Jelena GUČEVIĆ

    2013-09-01

    Full Text Available Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and some selected results.

  1. Calibration procedure for zenith plummets

    OpenAIRE

    Jelena GUČEVIĆ; Delčev, Siniša; Vukan OGRIZOVIĆ

    2013-01-01

    Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and som...

  2. Radiological Calibration and Standards Facility

    Data.gov (United States)

    Federal Laboratory Consortium — PNNL maintains a state-of-the-art Radiological Calibration and Standards Laboratory on the Hanford Site at Richland, Washington. Laboratory staff provide expertise...

  3. Calibration Techniques for VERITAS

    CERN Document Server

    Hanna, David

    2007-01-01

    VERITAS is an array of four identical telescopes designed for detecting and measuring astrophysical gamma rays with energies in excess of 100 GeV. Each telescope uses a 12 m diameter reflector to collect Cherenkov light from air showers initiated by incident gamma rays and direct it onto a `camera' comprising 499 photomultiplier tubes read out by flash ADCs. We describe here calibration methods used for determining the values of the parameters which are necessary for converting the digitized PMT pulses to gamma-ray energies and directions. Use of laser pulses to determine and monitor PMT gains is discussed, as are measurements of the absolute throughput of the telescopes using muon rings.

  4. TOD to TTP calibration

    Science.gov (United States)

    Bijl, Piet; Reynolds, Joseph P.; Vos, Wouter K.; Hogervorst, Maarten A.; Fanning, Jonathan D.

    2011-05-01

    The TTP (Targeting Task Performance) metric, developed at NVESD, is the current standard US Army model to predict EO/IR Target Acquisition performance. This model however does not have a corresponding lab or field test to empirically assess the performance of a camera system. The TOD (Triangle Orientation Discrimination) method, developed at TNO in The Netherlands, provides such a measurement. In this study, we make a direct comparison between TOD performance for a range of sensors and the extensive historical US observer performance database built to develop and calibrate the TTP metric. The US perception data were collected doing an identification task by military personnel on a standard 12 target, 12 aspect tactical vehicle image set that was processed through simulated sensors for which the most fundamental sensor parameters such as blur, sampling, spatial and temporal noise were varied. In the present study, we measured TOD sensor performance using exactly the same sensors processing a set of TOD triangle test patterns. The study shows that good overall agreement is obtained when the ratio between target characteristic size and TOD test pattern size at threshold equals 6.3. Note that this number is purely based on empirical data without any intermediate modeling. The calibration of the TOD to the TTP is highly beneficial to the sensor modeling and testing community for a variety of reasons. These include: i) a connection between requirement specification and acceptance testing, and ii) a very efficient method to quickly validate or extend the TTP range prediction model to new systems and tasks.

  5. Tectonic calibrations in molecular dating

    Institute of Scientific and Technical Information of China (English)

    Ullasa KODANDARAMAIAH

    2011-01-01

    Molecular dating techniques require the use of calibrations, which are usually fossil or geological vicariance-based.Fossil calibrations have been criticised because they result only in minimum age estimates. Based on a historical biogeographic perspective, Ⅰ suggest that vicariance-based calibrations are more dangerous. Almost all analytical methods in historical biogeography are strongly biased towards inferring vicariance, hence vicariance identified through such methods is unreliable. Other studies, especially of groups found on Gondwanan fragments, have simply assumed vicariance. Although it was previously believed that vicariance was the predominant mode of speciation, mounting evidence now indicates that speciation by dispersal is common, dominating vicariance in several groups. Moreover, the possibility of speciation having occurred before the said geological event cannot be precluded. Thus, geological calibrations can under- or overestimate times, whereas fossil calibrations always result in minimum estimates. Another major drawback of vicariant calibrations is the problem of circular reasoning when the resulting estimates are used to infer ages of biogeographic events. Ⅰ argue that fossil-based dating is a superior alternative to vicariance, primarily because the strongest assumption in the latter, that speciation was caused by the said geological process, is more often than not the most tenuous. When authors prefer to use a combination of fossil and vicariant calibrations, one suggestion is to report results both with and without inclusion of the geological constraints. Relying solely on vicariant calibrations should be strictly avoided.

  6. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Kristensen, L.; Jensen, G.; Hansen, A.;

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statisticalsignificance of the calibration expressions...

  7. Liquid Krypton Calorimeter Calibration Software

    CERN Document Server

    Hughes, Christina Lindsay

    2013-01-01

    Calibration of the liquid krypton calorimeter (LKr) of the NA62 experiment is managed by a set of standalone programs, or an online calibration driver. These programs are similar to those used by NA48, but have been updated to utilize classes and translated to C++ while maintaining a common functionality. A set of classes developed to handle communication with hardware was used to develop the three standalone programs as well as the main driver program for online calibration between bursts. The main calibration driver has been designed to respond to run control commands and receive burst data, both transmitted via DIM. In order to facilitate the process of reading in calibration parameters, a serializable class has been introduced, allowing the replacement of standard text files with XML configuration files.

  8. The Advanced LIGO Photon Calibrators

    CERN Document Server

    Karki, S; Kandhasamy, S; Abbott, B P; Abbott, T D; Anders, E H; Berliner, J; Betzwieser, J; Daveloza, H P; Cahillane, C; Canete, L; Conley, C; Gleason, J R; Goetz, E; Kissel, J S; Izumi, K; Mendell, G; Quetschke, V; Rodruck, M; Sachdev, S; Sadecki, T; Schwinberg, P B; Sottile, A; Wade, M; Weinstein, A J; West, M; Savage, R L

    2016-01-01

    The two interferometers of the Laser Interferometry Gravitaional-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events, and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as Photon Calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO Photon Calibrators that are currently providing fiducial displacements on the order of $10^{-18}$ m/$\\sqrt{\\textrm{Hz}}$ with accuracy and precision of better ...

  9. Antenna Calibration and Measurement Equipment

    Science.gov (United States)

    Rochblatt, David J.; Cortes, Manuel Vazquez

    2012-01-01

    A document describes the Antenna Calibration & Measurement Equipment (ACME) system that will provide the Deep Space Network (DSN) with instrumentation enabling a trained RF engineer at each complex to perform antenna calibration measurements and to generate antenna calibration data. This data includes continuous-scan auto-bore-based data acquisition with all-sky data gathering in support of 4th order pointing model generation requirements. Other data includes antenna subreflector focus, system noise temperature and tipping curves, antenna efficiency, reports system linearity, and instrument calibration. The ACME system design is based on the on-the-fly (OTF) mapping technique and architecture. ACME has contributed to the improved RF performance of the DSN by approximately a factor of two. It improved the pointing performances of the DSN antennas and productivity of its personnel and calibration engineers.

  10. TIME CALIBRATED OSCILLOSCOPE SWEEP CIRCUIT

    Science.gov (United States)

    Smith, V.L.; Carstensen, H.K.

    1959-11-24

    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  11. Requirements for gamma radiation survey meter calibration

    International Nuclear Information System (INIS)

    This guide describes the minimum requirements for calibrating a portable analog gamma radiation survey meter by means of a beam calibrator, with a known calibration source. If an alternative method of calibration is to be used the licensee should make a written request to the Atomic Energy Control Board that describes the calibration method to be used, and request the Board's permission to use that method in place of the requirements contained in this guide. This guide explains: the responsibility for survey meter calibration if licensees calibrate their own survey meters, use the services of a Canadian calibration agency, and use the services of a non-Canadian calibration agency; the requirements for survey meter calibration and the supporting documentation; the requirements for record-keeping; and, a calibration certificate, a calibration sticker, and a notification of failure to calibrate form, with examples

  12. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  13. del Nim

    Directory of Open Access Journals (Sweden)

    Arturo Martínez Rodríguez

    2007-01-01

    Full Text Available Dado su uso específico como fuente de insecticidas naturales de fácil biodegradación, entre otras ventajas, el cultivo del árbol Nim ha sido mundialmente extendido en muchas regiones de Asia, África, Australia y América Latina, incluyendo Cuba. La cosecha del fruto del Nim se hace engorrosa debido a la altura que adquieren estos árboles a los pocos años de sembrados, siendo la cosecha mecánica mediante sacudidores de ramas o del tronco, una posible vía de solución práctica de esta problemática. En el presente trabajo se efectúa, mediante la simulación con el empleo de herramientas CAD-3D y programas de análisis por elementos finitos, un análisis modal abarcador de un número importante de modos de vibración del sistema fruto-pedúnculo del Nim. Como resultado del estudio se pudieron recomendar entornos de frecuencias de las vibraciones a aplicar con mejores expectativas durante la cosecha mecanizada de este producto.

  14. Astrid-2 EMMA Magnetic Calibration

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Risbo, Torben;

    1998-01-01

    magnetometer readings in each position were related to the field magnitudes from the Observatory, and a least squares fit for the 9 magnetometer calibration parameters was performed (3 offsets, 3 scale values and 3 inter-axes angles). After corrections for the magnetometer digital-to-analogue converters...... fit calibration parameters. Owing to time shortage, we did not evaluate the temperature coefficients of the flight sensor calibration parameters. However, this was done for an identical flight spare magnetometer sensor at the magnetic coil facility belonging to the Technical University of Braunschweig...

  15. Potencial del agua del suelo

    OpenAIRE

    Bustamante Heliodoro

    2012-01-01

    La energía potencial del agua presenta diferencias de un punto del suelo a otro; esas diferencias son las que originan el movimiento del agua de acuerdo a la tendencia universal de la materia en el sentido de moverse de donde la energía potencial es mayor a donde dicha energía es menor. En el suelo el agua en consecuencia se mueve hacia donde su energía decrece hasta lograr su estado de equilibrio. Se desprende entonces que la cantidad de energía potencial absoluta contenida en el agua, no es...

  16. The calibration of radioprotection hardware

    International Nuclear Information System (INIS)

    After having recalled recent recommendations on dose limits on the basis of two radioprotection values (the equivalent and the efficient dose), this document indicates some characteristics of these values, and discusses how they are applied for individual monitoring and for area or ambient monitoring. It presents conventions aimed at simplifying radiation fields. Then, the author gives a precise overview of some general aspects concerning calibration operations: legal requirements, radioprotection hardware controls, calibration loop organisation (calibration definition, general physical values, reference radiation, conversion factors, and metrology), comparison between operational values and the protection value (irradiation geometries, conversion factors with respect to the geometries, comparison between efficient dose and operational values). He finally describes the calibration procedures: dosemeter location, energy response, angular response, flow rate response, uncertainties

  17. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... of non-uniformity of the illumination of the image plane. Only the image deforming aberrations and the non-uniformity of illumination are included in the calibration models. The topics of the pinhole camera model and the extension to the Direct Linear Transform (DLT) are described. It is shown how...... the correct interpolation method is described. For the chromatic issues of calibration we present the acquisition of colour and multi-spectral images, the chromatic aberrations and the various lens/camera based non-uniformities of the illumination of the image plane. It is described how the...

  18. Calibration of "Babyline" RP instruments

    CERN Document Server

    2015-01-01

      If you have old RP instrumentation of the “Babyline” type, as shown in the photo, please contact the Radiation Protection Group (Joffrey Germa, 73171) to have the instrument checked and calibrated. Thank you. Radiation Protection Group

  19. Rotary mode system initial instrument calibration

    Energy Technology Data Exchange (ETDEWEB)

    Johns, B.R.

    1994-10-01

    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files.

  20. Rotary mode system initial instrument calibration

    International Nuclear Information System (INIS)

    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files

  1. Beam Imaging and Luminosity Calibration

    CERN Document Server

    AUTHOR|(CDS)2081126; Klute, Markus; Medlock, Catherine Aiko

    2016-01-01

    We discuss a method to reconstruct two-dimensional proton bunch densities using vertex distributions accumulated during LHC beam-beam scans. The x-y correlations in the beam shapes are studied and an alternative luminosity calibration technique is introduced. We demonstrate the method on simulated beam-beam scans and estimate the uncertainty on the luminosity calibration associated to the beam-shape reconstruction to be below 1%.

  2. CERI: Ionizing Radiation Calibration Centre

    International Nuclear Information System (INIS)

    The CERI has been granted by the National Bureau of Metrology (BNM) as an Ionizing Radiation Calibration Centre and as an Estimation and Qualification Centre for the ionizing radiation measurement devices. This article gives some information on the scope covered by the BNM's grant and on the various equipment on which the laboratory relies. It describes the calibration and estimation activities and mentions many kinds of services which are offered to the users mainly in the medical and industrial fields

  3. The Third VLBA Calibrator Survey

    OpenAIRE

    Petrov, L.; Kovalev, Y. Y.; Fomalont, E.; Gordon, D

    2004-01-01

    This paper presents the third extension to the Very Large Baseline Array (VLBA) Calibrator Survey, containing 360 new sources not previously observed with very long baseline interferometry (VLBI). The survey, based on three 24 hour VLBA observing sessions, fills the areas on the sky above declination -45 degrees where the calibrator density is less than one source within a 4 degrees radius disk at any given direction. The positions were derived from astrometric analysis of the group delays de...

  4. Pressures Detector Calibration and Measurement

    CERN Document Server

    Kumara, I Made Gita

    2016-01-01

    This is report of my first and second projects (of 3) in NA61. I did data taking and analysis in order to do calibration of pressure detectors and verified it. I analyzed the data by ROOT software using the C ++ programming language. The first part of my project was determination of calibration factor of pressure sensors. Based on that result, I examined the relation between pressure drop, gas flow rate of in paper filter and its diameter.

  5. WFC3: UVIS Dark Calibration

    Science.gov (United States)

    Bourque, Matthew; Biretta, John A.; Anderson, Jay; Baggett, Sylvia M.; Gunning, Heather C.; MacKenty, John W.

    2014-06-01

    Wide Field Camera 3 (WFC3), a fourth-generation imaging instrument on board the Hubble Space Telescope (HST), has exhibited excellent performance since its installation during Servicing Mission 4 in May 2009. The UVIS detector, comprised of two e2v CCDs, is one of two channels available on WFC3 and is named for its ultraviolet and visible light sensitivity. We present the various procedures and results of the WFC3/UVIS dark calibration, which monitors the health and stability of the UVIS detector, provides characterization of hot pixels and dark current, and produces calibration files to be used as a correction for dark current in science images. We describe the long-term growth of hot pixels and the impacts that UVIS Charge Transfer Efficiency (CTE) losses, postflashing, and proximity to the readout amplifiers have on the population. We also discuss the evolution of the median dark current, which has been slowly increasing since the start of the mission and is currently ~6 e-/hr/pix, averaged across each chip. We outline the current algorithm for creating UVIS dark calibration files, which includes aggressive cosmic ray masking, image combination, and hot pixel flagging. Calibration products are available to the user community, typically 3-5 days after initial processing, through the Calibration Database System (CDBS). Finally, we discuss various improvements to the calibration and monitoring procedures. UVIS dark monitoring will continue throughout and beyond HST’s current proposal cycle.

  6. Calibration and Data Elaboration Procedure for Sky Irradiance Measurements

    Science.gov (United States)

    Boi, Paolo; Tonna, Glauco; Dalu, Giuseppe; Nakajima, Teruyuki; Olivieri, Bruno; Pompei, Alberto; Campanelli, Monica; Rao, R.

    1999-02-01

    The problems encountered in the elaboration of measurements of direct and sky diffuse solar irradiance are the following: (1) to carry out the calibration for the direct irradiance, which consists in determining the direct irradiance at the upper limit of the atmosphere; (2) to carry out the calibration for the diffuse irradiance, which consists in determining the solid viewing angle of the sky radiometer; (3) to determine the input parameters, namely, ground albedo, real and imaginary parts of the aerosol refractive index, and aerosol radius range; and (4) to determine from the optical data the columnar aerosol optical depth and volume radius distribution. With experimental data and numerical simulations a procedure is shown that enables one to carry out the two calibrations needed for the sky radiometer, to determine a best estimate of the input parameters, and, finally, to obtain the average features of the atmospheric aerosols. An interesting finding is that inversion of only data of diffuse irradiance yields the same accuracy of result as data of both diffuse and direct irradiance; in this case, only calibration of the solid viewing angle of the sky radiometer is needed, thus shortening the elaboration procedure. Measurements were carried out in the Western Mediterranean Sea (Italy), in Tokyo (Japan), and in Ushuaia (Tierra del Fuego, Argentina); data were elaborated with a new software package, the Skyrad code, based on an efficient radiative transfer scheme.

  7. Evaluation of quality assurance calibration results based on repeated calibrations

    International Nuclear Information System (INIS)

    To ensure quality assurance of the calibration results, as indicated by the UNE-EN ISO / IEC 17025:2005 in paragraph 5.9, the laboratory has established procedures for quality control of its activity. Thus, the laboratory participates in both inter-laboratory intercomparison exercises, cycle through the entire range of radiation qualities reflected in the scope of its accreditation, such as intra-laboratory intercomparison exercises. In this case, repeat quarterly by two different operators both the calibration of an ionization chamber irradiation of a direct reading personal dosimeter.

  8. The KLOE Online Calibration System

    Institute of Scientific and Technical Information of China (English)

    E.Pasqualucci

    2001-01-01

    Based on all the features of the KLOE online software,the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed.Acalibration manager process controls the system,implementing the interface to the online system,receiving information from the run control and translating its state transitions to a separate state machine.It acts as a " calibration run controller"and performs failure recovery when requested by a set of process checkers.The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms.A client library and C,fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an bool-like interface.Several calibration processes running in parallel in a destributed,multiplatform environment can fill the same histograms,allowing fast external information check.A monitor thread allow remote browsing for visual inspection,Pre-filtered data are read in nonprivileged spy mode from the data acquisition system via the Kloe Integrated Dataflow,privileged spy mode from the data acquisiton system via the Kole Integrated Dataflow.The main characteristics of the system are presented.

  9. Culturas del Mundo

    DEFF Research Database (Denmark)

    Benwell, Ann Fenger; Costa, Alberto; Waehle, Espen

    2006-01-01

    ’Culturas del mundo. Colecciones del Museo Nacional de Dinamarca’ with Ann Fenger Benwell in Culturas del Mundo. Colecciones del Museo Nacional de Dinamarca, ed. Silvia Sauquet, Fundación "la Caixa", Barcelona 2006, pp. 31-39......’Culturas del mundo. Colecciones del Museo Nacional de Dinamarca’ with Ann Fenger Benwell in Culturas del Mundo. Colecciones del Museo Nacional de Dinamarca, ed. Silvia Sauquet, Fundación "la Caixa", Barcelona 2006, pp. 31-39...

  10. The calibration system of the GERDA muon veto Cherenkov detector

    International Nuclear Information System (INIS)

    The GERDA experiment searches for neutrinoless double beta decay (0νββ). To achieve a sensitivity of 10-3counts/(keVkgy) or better within a specific region of interest (ROI), a good background identification is needed. Therefore GERDA is located in the LNGS (Laboratori Nationali del Gran Sasso) underground facility. In addition to the good rejection of cosmic muons due to the surrounding bedrocks, a dual muon veto system has to be used. For calibration and monitoring of the muon veto, two separate systems have been developed.

  11. Facility for dosimetric instrument calibration

    International Nuclear Information System (INIS)

    A structure is designed consisting of a rotary support of containers with radiation sources and of a rotary plug mounted above the sources. A support post with a slide rest and arms rotating around the post longitudinal axis is mounted in the centre of the container support. THe arms support instruments to be calibrated. The colimation cone of the respective source is directed to the tensor of the instrument being calibrated. The slide rest is balanced using a counterpoise mounted in the support post. Sources are not removed from the containers in source change during measurement. The support can hold more containers and the slide rest can support more instruments to be calibrated than the existing configurations. (M.D.). 2 figs

  12. Radiation protection dosimetry and calibrations

    International Nuclear Information System (INIS)

    At the SCK-CEN different specialised services are delivered for a whole range of external and internal customers in the radiation protection area. For the expertise group of radiation protection dosimetry and calibrations, these services are organized in four different laboratories: dosimetry, anthropogammametry, nuclear calibrations and non-nuclear calibrations. The services are given by a dedicated technical staff who has experience in the handling of routine and specialised cases. The scientific research that is performed by the expertise group makes sure that state-of-the-art techniques are being used, and that constant improvements and developments are implemented. Quality Assurance is an important aspect for the different services, and accreditation according national and international standards is achieved for all laboratories

  13. Optimal calibration of nuclear instrumentation

    International Nuclear Information System (INIS)

    Accurate knowledge of core power level is essential for the safe and efficient operation of nuclear power plants. Ionization chambers located outside the reactor core have the necessary reliability and response time characteristics and have been used extensively to indicate power level. The calibration of the ion chamber, and associated nuclear instrumentation (NI), has traditionally been based on the thermal power in the secondary coolant system. The usual NI calibration procedure consists of establishing steady-state operating conditions, calorimetrically determining the power at the secondary side of the steam generator, and adjusting the NI output to correspond to the measured thermal power. This study addresses certain questions including; (a) what sampling rate should be employed, (b) how many measurements are required, and (c) how can additional power level related information such as primary coolant loop measurements and knowledge of plant dynamics be included in the calibration procedure

  14. The Third VLBA Calibrator Survey

    CERN Document Server

    Petrov, L A; Fomalont, E; Gordon, D

    2005-01-01

    This paper presents the third extension to the Very Large Baseline Array (VLBA) Calibrator Survey, containing 360 new sources not previously observed with very long baseline interferometry (VLBI). The survey, based on three 24 hour VLBA observing sessions, fills the areas on the sky above declination -45 deg where the calibrator density is less than one source within a 4 deg radius disk at any given direction. The positions were derived from astrometric analysis of the group delays determined at 2.3 and 8.4 GHz frequency bands using the Calc/Solve package. The VCS3 catalogue of source positions, plots of correlated flux density versus the length of projected baseline, contour plots and fits files of naturally weighted CLEAN images as well as calibrated visibility function files are available on the Web at http://gemini.gsfc.nasa.gov/vcs3

  15. Power calibrations for TRIGA reactors

    International Nuclear Information System (INIS)

    The purpose of this paper is to establish a framework for the calorimetric power calibration of TRIGA reactors so that reliable results can be obtained with a precision better than ± 5%. Careful application of the same procedures has produced power calibration results that have been reproducible to ± 1.5%. The procedures are equally applicable to the Mark I, Mark II and Mark III reactors as well as to reactors having much larger reactor tanks and to TRIGA reactors capable of forced cooling up to 3 MW in some cases and 15 MW in another case. In the case of forced cooled TRIGA reactors, the calorimetric power calibration is applicable in the natural convection mode for these reactors using exactly the same procedures as are discussed below for the smaller TRIGA reactors (< 2 MW)

  16. Performance standard for dose Calibrator

    CERN Document Server

    Darmawati, S

    2002-01-01

    Dose calibrator is an instrument used in hospitals to determine the activity of radionuclide for nuclear medicine purposes. International Electrotechnical Commission (IEC) has published IEC 1303:1994 standard that can be used as guidance to test the performance of the instrument. This paper briefly describes content of the document,as well as explains the assessment that had been carried out to test the instrument accuracy in Indonesia through intercomparison measurement.Its is suggested that hospitals acquire a medical physicist to perform the test for its dose calibrator. The need for performance standard in the form of Indonesia Standard is also touched.

  17. Historia del experimento barométrico

    OpenAIRE

    Rosa María Herrera

    2012-01-01

    Ciertos problemas hidráulicos planteados por el ingeniero Baliani hacía 1630 culminaron en el experimento barométrico realizado por el matemático Torricelli en 1644. Este último, ferviente galileano, pensaba la matemática como lenguaje con que la naturaleza habla a los hombres. Con este soporte teórico del mundo realizó un experimento físico con el que calibró la presión atmosférica y comprobó la existencia del vacío.

  18. Historia del experimento barométrico

    Directory of Open Access Journals (Sweden)

    Rosa María Herrera

    2012-04-01

    Full Text Available Ciertos problemas hidráulicos planteados por el ingeniero Baliani hacía 1630 culminaron en el experimento barométrico realizado por el matemático Torricelli en 1644. Este último, ferviente galileano, pensaba la matemática como lenguaje con que la naturaleza habla a los hombres. Con este soporte teórico del mundo realizó un experimento físico con el que calibró la presión atmosférica y comprobó la existencia del vacío.

  19. Uniformity calibration for ICT image

    International Nuclear Information System (INIS)

    The uniformity of ICT image is impaired by beam hardening and the inconsistency of detector units responses. The beam hardening and the nonlinearity of the detector's output have been analyzed. The correction factors are determined experimentally by the detector's responses with different absorption length. The artifacts in the CT image of a symmetrical aluminium cylinder have been eliminated after calibration. (author)

  20. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Vesth, Allan; Kock, Carsten Weber

    The report describes power curve measurements carried out on a given wind turbine. The measurements are carried out in accordance to Ref. [1]. A site calibration has been carried out; see Ref. [2], and the measured flow correction factors for different wind directions are used in the present...

  1. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Villanueva, Héctor

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  2. Scalar Calibration of Vector Magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz;

    2000-01-01

    The calibration parameters of a vector magnetometer are estimated only by the use of a scalar reference magnetometer. The method presented in this paper differs from those previously reported in its linearized parametrization. This allows the determination of three offsets or signals in the absence...

  3. Autonomous Attitude Sensor Calibration (ASCAL)

    Science.gov (United States)

    Peterson, Chariya; Rowe, John; Mueller, Karl; Ziyad, Nigel

    1998-01-01

    In this paper, an approach to increase the degree of autonomy of flight software is proposed. We describe an enhancement of the Attitude Determination and Control System by augmenting it with self-calibration capability. Conventional attitude estimation and control algorithms are combined with higher level decision making and machine learning algorithms in order to deal with the uncertainty and complexity of the problem.

  4. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values...

  5. Recommended Inorganic Chemicals for Calibration.

    Science.gov (United States)

    Moody, John R.; And Others

    1988-01-01

    All analytical techniques depend on the use of calibration chemicals to relate analyte concentration to instrumental parameters. Discusses the preparation of standard solutions and provides a critical evaluation of available materials. Lists elements by group and discusses the purity and uses of each. (MVL)

  6. Calibration of personnel dose meters

    International Nuclear Information System (INIS)

    Methods of calibrating both film and thermoluminescent dose meters (TLD) to photon and electron radiations are described. K fluorescent X-rays, heavily filtered X-ray beams, and isotope gamma rays are used at the Los Alamos calibration facility to measure the energy and angular response of radiation detectors over a photon energy range of 10 to 1000keV. Beam spectra, alignment, size and uniformity are discussed. The energy and angular response of dose meters to electrons is measured with beta-emitting isotopes varying in maximum energy from 770 to 2300keV. A free-air ionization chamber is the primary standard used in the measurement of photon radiation. Thimble-sized ionization chambers, calibrated to the free-air chamber, serve as secondary standards. Electron radiation is measured with an end-window ionization chamber having a 7mg/cm2 approximately tissue-equivalent plastic wall. Photon calibrations are performed with personnel dose meters in air, on a phantom, and in a phantom. If the personnel dose meter and secondary chamber are both in air, or both on or both in a phantom, the response of the LiF TLD chip, relative to the secondary chamber, is the same. However, the film dose meter shows a larger relative response on or in the phantom than in air. With beta sources, personnel dose meters are calibrated by exposing the dose meter either in air to a high-dose-rate 90Sr (90Y) source, or in contact with a low-dose-rate uranium source. The differences in personnel dose meter response observed between the two methods are discussed. The personnel dose meters are calibrated to determine penetrating doses by placing the secondary chamber 1cm deep in a phantom and the personnel dose meter on the surface, with a filter over the TLD to simulate 1cm depth. Non-penetrating dose calibrations are measured by placing both chamber and dose meter on the surface of the phantom. (author)

  7. Mathematical efficiency calibration in gamma spectroscopy

    CERN Document Server

    Kaminski, S; Wilhelm, C

    2003-01-01

    Mathematical efficiency calibration with the LabSOCS software was introduced for two detectors in the measurement laboratory of the Central Safety Department of Forschungszentrum Karlsruhe. In the present contribution, conventional efficiency calibration of gamma spectroscopy systems and mathematical efficiency calibration with LabSOCS are compared with respect to their performance, uncertainties, expenses, and results. It is reported about the experience gained, and the advantages and disadvantages of both methods of efficiency calibration are listed. The results allow the conclusion to be drawn that mathematical efficiency calibration is a real alternative to conventional efficiency calibration of gamma spectroscopy systems as obtained by measurements of mixed gamma ray standard sources.

  8. Perfil del visitante del parque nacional Desierto del Carmen

    OpenAIRE

    Elizabeth López Carré

    2007-01-01

    El artículo presenta los resultados obtenidos del proyecto de investigación titulado "Evaluación del potencial de comercialización y caracterización del visitante del Santo Desierto de Tenancingo", específicamente los correspondientes al perfil del visitante a dicho Parque Nacional. La metodología utilizada se sustenta en la investigación de mercados propuesta por Kinnear y Taylor. El resultado de dicha investigación permitió integrar el perfil del segmento de mercado religioso...

  9. Sources of uncertainty in vicarious calibration: understanding calibration target reflectance

    OpenAIRE

    Anderson, K.; Milton, E. J.; Rollin, E.M.

    2003-01-01

    A field experiment investigated the hypothesis that the nadir reflectance of calibration surface substrates (asphalt and concrete) remains stable over a range of time-scales. Measurable differences in spectral reflectance factors were found over periods as short as 30 minutes. Multi-date reflectance measurements were compared using ANOVA and found to differ significantly (p = 0.001). Surface reflectance showed a relationship with the relative proportion of diffuse irradiance, over periods ...

  10. Extracting the MESA SR4000 calibrations

    Science.gov (United States)

    Charleston, Sean A.; Dorrington, Adrian A.; Streeter, Lee; Cree, Michael J.

    2015-05-01

    Time-of-flight range imaging cameras are capable of acquiring depth images of a scene. Some algorithms require these cameras to be run in `raw mode', where any calibrations from the off-the-shelf manufacturers are lost. The calibration of the MESA SR4000 is herein investigated, with an attempt to reconstruct the full calibration. Possession of the factory calibration enables calibrated data to be acquired and manipulated even in "raw mode." This work is motivated by the problem of motion correction, in which the calibration must be separated into component parts to be applied at different stages in the algorithm. There are also other applications, in which multiple frequencies are required, such as multipath interference correction. The other frequencies can be calibrated in a similar way, using the factory calibration as a base. A novel technique for capturing the calibration data is described; a retro-reflector is used on a moving platform, which acts as a point source at a distance, resulting in planar waves on the sensor. A number of calibrations are retrieved from the camera, and are then modelled and compared to the factory calibration. When comparing the factory calibration to both the "raw mode" data, and the calibration described herein, a root mean squared error improvement of 51:3mm was seen, with a standard deviation improvement of 34:9mm.

  11. Radiometric Calibration of Osmi Imagery Using Solar Calibration

    Science.gov (United States)

    Lee, Dong-Han; Kim, Yong-Seung

    2000-12-01

    OSMI (Ocean Scanning Multi-Spectral Imager) raw image data (Level 0) were acquired and radiometrically corrected. We have applied two methods, using solar & dark calibration data from OSMI sensor and comparing with the SeaWiFS data, to the radiometric correction of OSMI raw image data. First, we could get the values of the gain and the offset for each pixel and each band from comparing the solar & dark calibration data with the solar input radiance values, calculated from the transmittance, BRDF (Bidirectional Reflectance Distribution Function) and the solar incidence angle (¥â,¥è) of OSMI sensor. Applying this calibration data to OSMI raw image data, we got the two odd results, the lower value of the radiometric corrected image data than the expected value, and the Venetian Blind Effect in the radiometric corrected image data. Second, we could get the reasonable results from comparing OSMI raw image data with the SeaWiFS data, and get a new problem of OSMI sensor.

  12. Calibration and Validation of Measurement System

    DEFF Research Database (Denmark)

    Kofoed, Jens Peter; Riemann, Sven; Knapp, Wilfried

    The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype.......The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype....

  13. Attitude Sensor and Gyro Calibration for Messenger

    Science.gov (United States)

    O'Shaughnessy, Daniel; Pittelkau, Mark E.

    2007-01-01

    The Redundant Inertial Measurement Unit Attitude Determination/Calibration (RADICAL(TM)) filter was used to estimate star tracker and gyro calibration parameters using MESSENGER telemetry data from three calibration events. We present an overview of the MESSENGER attitude sensors and their configuration is given, the calibration maneuvers are described, the results are compared with previous calibrations, and variations and trends in the estimated calibration parameters are examined. The warm restart and covariance bump features of the RADICAL(TM) filter were used to estimate calibration parameters from two disjoint telemetry streams. Results show that the calibration parameters converge faster with much less transient variation during convergence than when the filter is cold-started at the start of each telemetry stream.

  14. Automated Calibration of Dosimeters for Diagnostic Radiology

    International Nuclear Information System (INIS)

    Calibration of dosimeters for diagnostic radiology includes current and charge measurements, which are often repetitive. However, these measurements are usually done using modern electrometers, which are equipped with an RS-232 interface that enables instrument control from a computer. This paper presents an automated system aimed to the measurements for the calibration of dosimeters used in diagnostic radiology. A software application was developed, in order to achieve the acquisition of the electric charge readings, measured values of the monitor chamber, calculation of the calibration coefficient and issue of a calibration certificate. A primary data record file is filled and stored in the computer hard disk. The calibration method used was calibration by substitution. With this system, a better control over the calibration process is achieved and the need for human intervention is reduced. the automated system will be used in the calibration of dosimeters for diagnostic radiology at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene. (Author)

  15. Automatic calibration system for pressure transducers

    Science.gov (United States)

    1968-01-01

    Fifty-channel automatic pressure transducer calibration system increases quantity and accuracy for test evaluation calibration. The pressure transducers are installed in an environmental tests chamber and manifolded to connect them to a pressure balance which is uniform.

  16. Quality control for dose calibrators

    International Nuclear Information System (INIS)

    Nuclear medicine laboratories are required to assay samples of radioactivity to be administered to patients. Almost universally, these assays are accomplished by use of a well ionization chamber isotope calibrator. The Instituto de Radioprotecao e Dosimetria (Institute for Radiological Protection and Dosimetry) of the Comissao Nacional de Energia Nuclear (National Commission for Nuclear Energy) is carrying out a National Quality Control Programme in Nuclear Medicine, supported by the International Atomic Energy Agency. The assessment of the current needs and practices of quality control in the entire country of Brazil includes Dose Calibrators and Scintillation Cameras, but this manual is restricted to the former. Quality Control Procedures for these Instruments are described in this document together with specific recommendations and assessment of its accuracy. (author)

  17. Spinning angle optical calibration apparatus

    Science.gov (United States)

    Beer, Stephen K.; Pratt, II, Harold R.

    1991-01-01

    An optical calibration apparatus is provided for calibrating and reproducing spinning angles in cross-polarization, nuclear magnetic resonance spectroscopy. An illuminated magnifying apparatus enables optical setting an accurate reproducing of spinning "magic angles" in cross-polarization, nuclear magnetic resonance spectroscopy experiments. A reference mark scribed on an edge of a spinning angle test sample holder is illuminated by a light source and viewed through a magnifying scope. When the "magic angle" of a sample material used as a standard is attained by varying the angular position of the sample holder, the coordinate position of the reference mark relative to a graduation or graduations on a reticle in the magnifying scope is noted. Thereafter, the spinning "magic angle" of a test material having similar nuclear properties to the standard is attained by returning the sample holder back to the originally noted coordinate position.

  18. Sky camera geometric calibration using solar observations

    OpenAIRE

    Urquhart, B.; Kurtz, B; J. Kleissl

    2016-01-01

    A camera model and associated automated calibration procedure for stationary daytime sky imaging cameras is presented. The specific modeling and calibration needs are motivated by remotely deployed cameras used to forecast solar power production where cameras point skyward and use 180° fisheye lenses. Sun position in the sky and on the image plane provides a simple and automated approach to calibration; special equipment or calibration patterns are not required. Sun positio...

  19. Binary Classifier Calibration: Non-parametric approach

    OpenAIRE

    Naeini, Mahdi Pakdaman; Cooper, Gregory F.; Hauskrecht, Milos

    2014-01-01

    Accurate calibration of probabilistic predictive models learned is critical for many practical prediction and decision-making tasks. There are two main categories of methods for building calibrated classifiers. One approach is to develop methods for learning probabilistic models that are well-calibrated, ab initio. The other approach is to use some post-processing methods for transforming the output of a classifier to be well calibrated, as for example histogram binning, Platt scaling, and is...

  20. WCDMA User Equipment Output Power Calibration

    OpenAIRE

    Folkeson, Tea

    2003-01-01

    To save time in Flextronics high volume production, the time for test and calibration of mobile telephones need to be as short and accurate as possible. In the wideband code division multiple access (WCDMA) case, the output power calibration is the most critical calibration concerning accuracy. The aim with this thesis was to find a faster calibration method than the one that exists today and still retain accuracy. The Third Generation Partnership Project (3GPP) outlines the requirements of ...

  1. GREAT/SOFIA atmospheric calibration

    OpenAIRE

    Guan, Xin; Stutzki, Jürgen; Graf, Urs U.; Güsten, Rolf; Okada, Yoko; Torres, Miguel Angel Requena; Simon, Robert; Wiesemeyer, Helmut

    2012-01-01

    The GREAT observations need frequency-selective calibration across the passband for the residual atmospheric opacity at flight altitude. At these altitudes the atmospheric opacity has both narrow and broad spectral features. To determine the atmospheric transmission at high spectral resolution, GREAT compares the observed atmospheric emission with atmospheric model predictions, and therefore depends on the validity of the atmospheric models. We discusse the problems identified in this compari...

  2. Automated Camera Array Fine Calibration

    Science.gov (United States)

    Clouse, Daniel; Padgett, Curtis; Ansar, Adnan; Cheng, Yang

    2008-01-01

    Using aerial imagery, the JPL FineCalibration (JPL FineCal) software automatically tunes a set of existing CAHVOR camera models for an array of cameras. The software finds matching features in the overlap region between images from adjacent cameras, and uses these features to refine the camera models. It is not necessary to take special imagery of a known target and no surveying is required. JPL FineCal was developed for use with an aerial, persistent surveillance platform.

  3. Colour calibration for colour reproduction

    OpenAIRE

    Emmel, P.; R. D. Hersch

    2000-01-01

    Due to the proliferation of low-cost colour devices (digital colour cameras, scanners, printers etc.) during the last few years, colour calibration has become an important issue. Such devices should faithfully reproduce colour images, but experience shows they don't. Among the main reasons, we note the diversity of acquisition, display and printing technologies which makes standardization difficult. Each device has a different gamut, i.e. a different set of colours that it can acquire or repr...

  4. Multidetector calibration for mass spectrometers

    International Nuclear Information System (INIS)

    The International Atomic Energy Agency's Safeguards Analytical Laboratory has performed calibration experiments to measure the different efficiencies among multi-Faraday detectors for a Finnigan-MAT 261 mass spectrometer. Two types of calibration experiments were performed: (1) peak-shift experiments and (2) peak-jump experiments. For peak-shift experiments, the ion intensities were measured for all isotopes of an element in different Faraday detectors. Repeated measurements were made by shifting the isotopes to various Faraday detectors. Two different peak-shifting schemes were used to measure plutonium (UK Pu5/92138) samples. For peak-jump experiments, ion intensities were measured in a reference Faraday detector for a single isotope and compared with those measured in the other Faraday detectors. Repeated measurements were made by switching back-and-forth between the reference Faraday detector and a selected Faraday detector. This switching procedure is repeated for all Faraday detectors. Peak-jump experiments were performed with replicate measurements of 239Pu, 187Re, and 238U. Detector efficiency factors were estimated for both peak-jump and peak-shift experiments using a flexible calibration model to statistically analyze both types of multidetector calibration experiments. Calculated detector efficiency factors were shown to depend on both the material analyzed and the experimental conditions. A single detector efficiency factor is not recommended for each detector that would be used to correct routine sample analyses. An alternative three-run peak-shift sample analysis should be considered. A statistical analysis of the data from this peak-shift experiment can adjust the isotopic ratio estimates for detector differences due to each sample analysis

  5. ASCAL: Autonomous Attitude Sensor Calibration

    Science.gov (United States)

    Peterson, Chariya; Rowe, John; Mueller, Karl; Ziyad, Nigel

    1999-01-01

    Abstract In this paper, an approach to increase the degree of autonomy of flight software is proposed. We describe an enhancement of the Attitude Determination and Control System by augmenting it with self-calibration capability. Conventional attitude estimation and control algorithms are combined with higher level decision making and machine learning algorithms in order to deal with the uncertainty and complexity of the problem.

  6. Calibrating thermal behavior of electronics

    Science.gov (United States)

    Chainer, Timothy J.; Parida, Pritish R.; Schultz, Mark D.

    2016-05-31

    A method includes determining a relationship between indirect thermal data for a processor and a measured temperature associated with the processor, during a calibration process, obtaining the indirect thermal data for the processor during actual operation of the processor, and determining an actual significant temperature associated with the processor during the actual operation using the indirect thermal data for the processor during actual operation of the processor and the relationship.

  7. PACS photometer calibration block analysis

    CERN Document Server

    Moór, A; Kiss, Cs; Balog, Z; Billot, N; Marton, G

    2013-01-01

    The absolute stability of the PACS bolometer response over the entire mission lifetime without applying any corrections is about 0.5% (standard deviation) or about 8% peak-to-peak. This fantastic stability allows us to calibrate all scientific measurements by a fixed and time-independent response file, without using any information from the PACS internal calibration sources. However, the analysis of calibration block observations revealed clear correlations of the internal source signals with the evaporator temperature and a signal drift during the first half hour after the cooler recycling. These effects are small, but can be seen in repeated measurements of standard stars. From our analysis we established corrections for both effects which push the stability of the PACS bolometer response to about 0.2% (stdev) or 2% in the blue, 3% in the green and 5% in the red channel (peak-to-peak). After both corrections we still see a correlation of the signals with PACS FPU temperatures, possibly caused by parasitic h...

  8. Unassisted 3D camera calibration

    Science.gov (United States)

    Atanassov, Kalin; Ramachandra, Vikas; Nash, James; Goma, Sergio R.

    2012-03-01

    With the rapid growth of 3D technology, 3D image capture has become a critical part of the 3D feature set on mobile phones. 3D image quality is affected by the scene geometry as well as on-the-device processing. An automatic 3D system usually assumes known camera poses accomplished by factory calibration using a special chart. In real life settings, pose parameters estimated by factory calibration can be negatively impacted by movements of the lens barrel due to shaking, focusing, or camera drop. If any of these factors displaces the optical axes of either or both cameras, vertical disparity might exceed the maximum tolerable margin and the 3D user may experience eye strain or headaches. To make 3D capture more practical, one needs to consider unassisted (on arbitrary scenes) calibration. In this paper, we propose an algorithm that relies on detection and matching of keypoints between left and right images. Frames containing erroneous matches, along with frames with insufficiently rich keypoint constellations, are detected and discarded. Roll, pitch yaw , and scale differences between left and right frames are then estimated. The algorithm performance is evaluated in terms of the remaining vertical disparity as compared to the maximum tolerable vertical disparity.

  9. Model Calibration in Watershed Hydrology

    Science.gov (United States)

    Yilmaz, Koray K.; Vrugt, Jasper A.; Gupta, Hoshin V.; Sorooshian, Soroosh

    2009-01-01

    Hydrologic models use relatively simple mathematical equations to conceptualize and aggregate the complex, spatially distributed, and highly interrelated water, energy, and vegetation processes in a watershed. A consequence of process aggregation is that the model parameters often do not represent directly measurable entities and must, therefore, be estimated using measurements of the system inputs and outputs. During this process, known as model calibration, the parameters are adjusted so that the behavior of the model approximates, as closely and consistently as possible, the observed response of the hydrologic system over some historical period of time. This Chapter reviews the current state-of-the-art of model calibration in watershed hydrology with special emphasis on our own contributions in the last few decades. We discuss the historical background that has led to current perspectives, and review different approaches for manual and automatic single- and multi-objective parameter estimation. In particular, we highlight the recent developments in the calibration of distributed hydrologic models using parameter dimensionality reduction sampling, parameter regularization and parallel computing.

  10. Calibration of RB reactor power

    International Nuclear Information System (INIS)

    The first and only calibration of RB reactor power was done in 1962, and the obtained calibration ratio was used irrespective of the lattice pitch and core configuration. Since the RB reactor is being prepared for operation at higher power levels it was indispensable to reexamine the calibration ratio, estimate its dependence on the lattice pitch, critical level of heavy water and thickness of the side reflector. It was necessary to verify the reliability of control and dosimetry instruments, and establish neutron and gamma dose dependence on reactor power. Two series of experiments were done in June 1976. First series was devoted to tests of control and dosimetry instrumentation and measurements of radiation in the RB reactor building dependent on reactor power. Second series covered measurement of thermal and epithermal neuron fluxes in the reactor core and calculation of reactor power. Four different reactor cores were chosen for these experiments. Reactor pitches were 8, 8√2, and 16 cm with 40, 52 and 82 fuel channels containing 2% enriched fuel. Obtained results and analysis of these results are presented in this document with conclusions related to reactor safe operation

  11. A Numerical Simulator for Noise Calibration Studies

    OpenAIRE

    Racette, Paul; Clune, Tom; Wong, Mark; Walker, David; Coakley, Kevin; Splett, Jolene; Rivers, Derick; Leonard, Robert; Boone, Ed

    2012-01-01

    Frequent calibration using noise references is used to reduce the effects of time varying fluctuations that naturally occur within sensitive radiometer receivers. Over the years, calibration architectures and processing algorithms have become more sophisticated. Predicting the performance of a given calibration architecture often requires expensive prototyping of hardware; optimizing processing algorithms for nonstationary fluctuations is a challenge. Measurement uncertainty is a figure of me...

  12. 48 CFR 908.7113 - Calibration services.

    Science.gov (United States)

    2010-10-01

    ... 48 Federal Acquisition Regulations System 5 2010-10-01 2010-10-01 false Calibration services. 908... ACQUISITION PLANNING REQUIRED SOURCES OF SUPPLIES AND SERVICES Acquisition of Special Items 908.7113 Calibration services. Orders for calibration services may be placed with the National Institute of...

  13. MODIS Radiometric Calibration and Uncertainty Assessment

    Science.gov (United States)

    Xiong, Xiaoxiong; Chiang, Vincent; Sun, Junqiang; Wu, Aisheng

    2011-01-01

    Since launch, Terra and Aqua MODIS have collected more than II and 9 years of datasets for comprehensive studies of the Earth's land, ocean, and atmospheric properties. MODIS observations are made in 36 spectral bands: 20 reflective solar bands (RSB) and 16 thermal emissive bands (TEB). Compared to its heritage sensors, MODIS was developed with very stringent calibration and uncertainty requirements. As a result, MODIS was designed and built with a set of state of the art on-board calibrators (OBC), which allow key sensor performance parameters and on-orbit calibration coefficients to be monitored and updated if necessary. In terms of its calibration traceability, MODIS RSB calibration is reflectance based using an on-board solar diffuser (SD) and the TEB calibration is radiance based using an on-board blackbody (BB). In addition to on-orbit calibration coefficients derived from its OBC, calibration parameters determined from sensor pre-launch calibration and characterization are used in both the RSB and TEB calibration and retrieval algorithms. This paper provides a brief description of MODIS calibration methodologies and discusses details of its on-orbit calibration uncertainties. It assesses uncertainty contributions from individual components and differences between Terra and Aqua MODIS due to their design characteristics and on-orbit periormance. Also discussed in this paper is the use of MODIS LIB uncertainty index CUI) product.

  14. New method for calibration of sun photometers

    Institute of Scientific and Technical Information of China (English)

    H.H.Asadov; I.G.Chobanzadeh

    2009-01-01

    A new method for calibration of sun photometers based on Bouguer-Beer law is proposed.The developed basic equation of calibration makes it possible to formulate the derivative methods of calibration on the basis of photometric measurements upon optical air masses,the ratio of which is an integer number.

  15. Net analyte signal calculation for multivariate calibration

    NARCIS (Netherlands)

    Ferre, J.; Faber, N.M.

    2003-01-01

    A unifying framework for calibration and prediction in multivariate calibration is shown based on the concept of the net analyte signal (NAS). From this perspective, the calibration step can be regarded as the calculation of a net sensitivity vector, whose length is the amount of net signal when the

  16. Calibration of nuclear medicine gamma counters

    International Nuclear Information System (INIS)

    In this paper the practical problem of nuclear medicine gamma counters calibration has been solved by using dose calibrators CRC-15R with standard error ±5%. The samples from technetium generators have been measured both by dose calibrators CRC-15R and gamma counter ICN Gamma 3.33 taking into account decay correction. Only the linear part of the curve has practical meaning. The advantage of this procedure satisfies the requirements from international standards: the calibration of sources used for medical exposure be traceable to a standard dosimetry laboratory and radiopharmaceuticals for nuclear medicine procedures be calibrated in terms of activity of the radiopharmaceutical to be administered. (author)

  17. Borexino calibrations: Hardware, Methods, and Results

    CERN Document Server

    Back, H; Benziger, J; Bick, D; Bonfini, G; Bravo, D; Avanzini, M Buizza; Caccianiga, B; Cadonati, L; Calaprice, F; Carraro, C; Cavalcante, P; Chavarria, A; D'Angelo, D; Davini, S; Derbin, A; Etenko, A; von Feilitzsch, F; Fernandes, G; Fomenko, K; Franco, D; Galbiati, C; Gazzana, S; Ghiano, C; Giammarchi, M; Goeger-Neff, M; Goretti, A; Grandi, L; Guardincerri, E; Hardy, S; Ianni, Aldo; Ianni, Andrea; Kayunov, A; Kidner, S; Kobychev, V; Korablev, D; Korga, G; Koshio, Y; Kryn, D; Laubenstein, M; Lewke, T; Litvinovich, E; Loer, B; Lombardi, F; Lombardi, P; Ludhova, L; Machulin, I; Manecki, S; Maneschg, W; Manuzio, G; Meindl, Q; Meroni, E; Miramonti, L; Misiaszek, M; Montanari, D; Mosteiro, P; Muratova, V; Oberauer, L; Obolensky, M; Ortica, F; Otis, K; Pallavicini, M; Papp, L; Perasso, L; Perasso, S; Pocar, A; Raghavan, R S; Ranucci, G; Razeto, A; Re, A; Romani, A; Rossi, N; Rountree, D; Sabelnikov, A; Saldanha, R; Salvo, C; Schonert, S; Simgen, H; Skorokhvatov, M; Smirnov, O; Sotnikov, A; Sukhotin, S; Suvorov, Y; Tartaglia, R; Testera, G; Vignaud, D; Vogelaar, R B; Winter, J; Wojcik, M; Wright, A; Wurm, M; Xu, J; Zaimidoroga, O; Zavatarelli, S; Zuzel, G

    2012-01-01

    Borexino was the first experiment to detect solar neutrinos in real-time in the sub-MeV region. In order to achieve high precision in the determination of neutrino rates, the detector design includes an internal and an external calibration system. This paper describes both calibration systems and the calibration campaigns that were carried out in the period between 2008 and 2011. We discuss some of the results and show that the calibration procedures preserved the radiopurity of the scintillator. The calibrations provided a detailed understanding of the detector response and led to a significant reduction of the systematic uncertainties in the Borexino measurements.

  18. Calibration of detector sensitivity in positron cameras

    International Nuclear Information System (INIS)

    An improved method for calibrating detector sensitivities in a positron camera has been developed. The calibration phantom is a cylinder of activity placed near the center of the camera and fully within the field of view. The calibration data is processed in such a manner that the following two important properties are achieved. The estimate of a detector sensitivity is unaffected by the sensitivities of the other detectors. The estimates are insensitive to displacements of the calibrating phantom from the camera center. Both of these properties produce a more accurate detector calibration

  19. ATLAS calorimeters energy calibration for jets

    International Nuclear Information System (INIS)

    The calibration of ATLAS barrel calorimeters (including pre shower system, electromagnetic Liquid Argon calorimeter and scintillating hadron tile calorimeter) was done by standard calibration and weighting technique approaches. The standard calibration gives the bad linearity for hadron non compensated calorimeter. The calibration with weighting technique, in comparison with standard calibration, restores linearity and improves energy resolution up to (σ/E)2 = (38.6%/√E)2 + (1.5%)2 for η 0.6. 6 refs., 4 figs., 1 tab

  20. Muon Energy Calibration of the MINOS Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Miyagawa, Paul S.

    2004-09-01

    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by {approx} 10%, which is equivalent to increasing the amount of data by 20%.

  1. MODIS Instrument Operation and Calibration Improvements

    Science.gov (United States)

    Xiong, X.; Angal, A.; Madhavan, S.; Link, D.; Geng, X.; Wenny, B.; Wu, A.; Chen, H.; Salomonson, V.

    2014-01-01

    Terra and Aqua MODIS have successfully operated for over 14 and 12 years since their respective launches in 1999 and 2002. The MODIS on-orbit calibration is performed using a set of on-board calibrators, which include a solar diffuser for calibrating the reflective solar bands (RSB) and a blackbody for the thermal emissive bands (TEB). On-orbit changes in the sensor responses as well as key performance parameters are monitored using the measurements of these on-board calibrators. This paper provides an overview of MODIS on-orbit operation and calibration activities, and instrument long-term performance. It presents a brief summary of the calibration enhancements made in the latest MODIS data collection 6 (C6). Future improvements in the MODIS calibration and their potential applications to the S-NPP VIIRS are also discussed.

  2. Calibration of the Cherenkov Telescope Array

    CERN Document Server

    Gaug, Markus; Berge, David; Reyes, Raquel de los; Doro, Michele; Foerster, Andreas; Maccarone, Maria Concetta; Parsons, Dan; van Eldik, Christopher

    2015-01-01

    The construction of the Cherenkov Telescope Array is expected to start soon. We will present the baseline methods and their extensions currently foreseen to calibrate the observatory. These are bound to achieve the strong requirements on allowed systematic uncertainties for the reconstructed gamma-ray energy and flux scales, as well as on the pointing resolution, and on the overall duty cycle of the observatory. Onsite calibration activities are designed to include a robust and efficient calibration of the telescope cameras, and various methods and instruments to achieve calibration of the overall optical throughput of each telescope, leading to both inter-telescope calibration and an absolute calibration of the entire observatory. One important aspect of the onsite calibration is a correct understanding of the atmosphere above the telescopes, which constitutes the calorimeter of this detection technique. It is planned to be constantly monitored with state-of-the-art instruments to obtain a full molecular and...

  3. Multi-Axis Accelerometer Calibration System

    Science.gov (United States)

    Finley, Tom; Parker, Peter

    2010-01-01

    A low-cost, portable, and simplified system has been developed that is suitable for in-situ calibration and/or evaluation of multi-axis inertial measurement instruments. This system overcomes facility restrictions and maintains or improves the calibration quality for users of accelerometer-based instruments with applications in avionics, experimental wind tunnel research, and force balance calibration applications. The apparatus quickly and easily positions a multi-axis accelerometer system into a precisely known orientation suitable for in-situ quality checks and calibration. In addition, the system incorporates powerful and sophisticated statistical methods, known as response surface methodology and statistical quality control. These methods improve calibration quality, reduce calibration time, and allow for increased calibration frequency, which enables the monitoring of instrument stability over time.

  4. Laser Calibration of an Impact Disdrometer

    Science.gov (United States)

    Lane, John E.; Kasparis, Takis; Metzger, Philip T.; Jones, W. Linwood

    2014-01-01

    A practical approach to developing an operational low-cost disdrometer hinges on implementing an effective in situ adaptive calibration strategy. This calibration strategy lowers the cost of the device and provides a method to guarantee continued automatic calibration. In previous work, a collocated tipping bucket rain gauge was utilized to provide a calibration signal to the disdrometer's digital signal processing software. Rainfall rate is proportional to the 11/3 moment of the drop size distribution (a 7/2 moment can also be assumed, depending on the choice of terminal velocity relationship). In the previous case, the disdrometer calibration was characterized and weighted to the 11/3 moment of the drop size distribution (DSD). Optical extinction by rainfall is proportional to the 2nd moment of the DSD. Using visible laser light as a means to focus and generate an auxiliary calibration signal, the adaptive calibration processing is significantly improved.

  5. The Palomar Testbed Interferometer Calibrator Catalog

    CERN Document Server

    Van Belle, G T; Creech-Eakman, M J; Coyne, J; Boden, A F; Akeson, R L; Ciardi, D R; Rykoski, K M; Thompson, R R; Lane, B F

    2007-01-01

    The Palomar Testbed Interferometer (PTI) archive of observations between 1998 and 2005 is examined for objects appropriate for calibration of optical long-baseline interferometer observations - stars that are predictably point-like and single. Approximately 1,400 nights of data on 1,800 objects were examined for this investigation. We compare those observations to an intensively studied object that is a suitable calibrator, HD217014, and statistically compare each candidate calibrator to that object by computing both a Mahalanobis distance and a Principal Component Analysis. Our hypothesis is that the frequency distribution of visibility data associated with calibrator stars differs from non-calibrator stars such as binary stars. Spectroscopic binaries resolved by PTI, objects known to be unsuitable for calibrator use, are similarly tested to establish detection limits of this approach. From this investigation, we find more than 350 observed stars suitable for use as calibrators (with an additional $\\approx 1...

  6. Pulse-based internal calibration of polarimetric SAR

    DEFF Research Database (Denmark)

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel...

  7. Protocols for calibrating multibeam sonar

    OpenAIRE

    Foote, Kenneth G.; Chu, Dezhang; Hammar, Terence R.; Baldwin, Kenneth C.; Mayer, Larry A.; Hufnagle, Lawrence C. jr.; Jech, J. Michael

    2005-01-01

    Development of protocols for calibrating multibeam sonar by means of the standard-target method is documented. Particular systems used in the development work included three that provide the water-column signals, namely the SIMRAD SM2000/90- and 200-kHz sonars and RESON SeaBat 8101 sonar, with operating frequency of 240 kHz. Two facilities were instrumented specifically for the work: a sea well at the Woods Hole Oceanographic Institution and a large, indoor freshwater tank at the University o...

  8. Calibration of TOB+ Thermometer's Cards

    CERN Document Server

    Banitt, Daniel

    2014-01-01

    Motivation - Under the new upgrade of the CMS detector the working temperature of the trackers had been reduced to -27 Celsius degrees. Though the thermal sensors themselves (Murata and Fenwal thermistors) are effective at these temperatures, the max1542 PLC (programmable logic controller) cards, interpreting the resistance of the thermal sensors into DC counts usable by the DCS (detector control system), are not designed for these temperatures in which the counts exceed their saturation and therefor had to be replaced. In my project I was in charge of handling the emplacement and calibration of the new PLC cards to the TOB (tracker outer barrel) control system.

  9. Scintillation camera brightness calibration apparatus

    International Nuclear Information System (INIS)

    Circuitry is described for calibrating the brightness of a cathode ray tube display and recording apparatus comprising: 1) intensity control means for adjusting the intensity of the cathode ray tube beam; 2) light sensitive means disposed to receive light emitted from the cathode ray tube and generating a first electrical signal having a magnitude dependent upon the intensity of the emitted light; 3) reference signal generating means for generating a second electrical signal of predetermined magnitude; and 4) electrical signal comparison means coupled to the light sensitive means and the reference signal generating means for comparing the magnitude of the first and second electrical signals. (author)

  10. Scalar Calibration of Vector Magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz; Petersen, Jan Raagaard; Nielsen, Otto V

    2000-01-01

    uniquely the parameters for a given data set. Therefore, a magnetometer may be characterized inexpensively in the Earth's magnetic-field environment. This procedure has been used successfully in the pre-flight calibration of the state-of-the-art magnetometers on board the magnetic mapping satellites Orsted...... of a magnetic field, three scale factors for normalization of the axes and three non-orthogonality angles which build up an orthogonal system intrinsically in the sensor. The advantage of this method compared with others lies in its linear least squares estimator, which finds independently and...

  11. Percutaneous transhepatic fine calibre cholangioscopy

    Energy Technology Data Exchange (ETDEWEB)

    Klose, K.J.; Thelen, M.; Schild, H.H.

    1988-07-01

    Percutaneous transhepatic fine calibre cholangioscopy is described. A specially developed transparent instrument is used, which makes it unnecessary to have a steerable endoscope; visualisation of the biliary tree is of diagnostic value for assessing biliary stenoses due to tumours (choice of biopsy site, extent of intraductal radiation therapy) and for the assessment of anastomoses between the biliary system and the gut (condition of the mucosa, stenoses). In the presence of percutaneous biliary drainage, the method has little utility. Extension of the method for treating intraductal stones may be possible if in future a useful lumen can be added to the fine endoscope.

  12. del cambio

    Directory of Open Access Journals (Sweden)

    Alejandro Canales

    2006-01-01

    Full Text Available El artículo analiza las propuestas y los compromisos en materia de cobertura y de financiamiento en el periodo 2000-2006. En particular, revisa las modificaciones y los ajustes de los tres principales documentos en los que se expresaron las iniciativas: el de la Asociación Nacional de Universidades e Instituciones de Educación Superior, el del equipo de la transición y el programa sectorial. Los resultados muestran que las cifras de atención a la demanda educativa e inversión quedaron lejos de las metas autoimpuestas por el gobierno federal y que será necesario realizar mayores esfuerzos para atender tales desafíos.

  13. MODIS Solar Reflective Calibration Traceability

    Science.gov (United States)

    Xiong, Xiaoxiong; Butler, Jim

    2009-01-01

    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify, measurement uncertainties, and to establish absolute scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2 micrometers and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bidirectional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser monitor (SDSM). This paper provides details of this calibration chain, from prelaunch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  14. MODIS solar reflective calibration traceability

    Science.gov (United States)

    Xiong, Xiaoxiong; Butler, Jim

    2009-08-01

    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify measurement uncertainties, and to establish an absolute measurement scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2μm and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bi-directional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser stability monitor (SDSM). This paper provides details of this calibration chain, from pre-launch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  15. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    Science.gov (United States)

    Cornic, Philippe; Illoul, Cédric; Cheminet, Adam; Le Besnerais, Guy; Champagnat, Frédéric; Le Sant, Yves; Leclaire, Benjamin

    2016-09-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data.

  16. Herschel SPIRE FTS Relative Spectral Response Calibration

    CERN Document Server

    Fulton, Trevor; Baluteau, Jean-Paul; Benielli, Dominique; Imhof, Peter; Lim, Tanya; Lu, Nanyao; Marchili, Nicola; Naylor, David; Polehampton, Edward; Swinyard, Bruce; Valtchanov, Ivan

    2014-01-01

    Herschel/SPIRE Fourier transform spectrometer (FTS) observations contain emission from both the Herschel Telescope and the SPIRE Instrument itself, both of which are typically orders of magnitude greater than the emission from the astronomical source, and must be removed in order to recover the source spectrum. The effects of the Herschel Telescope and the SPIRE Instrument are removed during data reduction using relative spectral response calibration curves and emission models. We present the evolution of the methods used to derive the relative spectral response calibration curves for the SPIRE FTS. The relationship between the calibration curves and the ultimate sensitivity of calibrated SPIRE FTS data is discussed and the results from the derivation methods are compared. These comparisons show that the latest derivation methods result in calibration curves that impart a factor of between 2 and 100 less noise to the overall error budget, which results in calibrated spectra for individual observations whose n...

  17. Absolute calibration in vivo measurement systems

    International Nuclear Information System (INIS)

    Lawrence Livermore National Laboratory (LLNL) is currently investigating a new method for obtaining absolute calibration factors for radiation measurement systems used to measure internally deposited radionuclides in vivo. Absolute calibration of in vivo measurement systems will eliminate the need to generate a series of human surrogate structures (i.e., phantoms) for calibrating in vivo measurement systems. The absolute calibration of in vivo measurement systems utilizes magnetic resonance imaging (MRI) to define physiological structure, size, and composition. The MRI image provides a digitized representation of the physiological structure, which allows for any mathematical distribution of radionuclides within the body. Using Monte Carlo transport codes, the emission spectrum from the body is predicted. The in vivo measurement equipment is calibrated using the Monte Carlo code and adjusting for the intrinsic properties of the detection system. The calibration factors are verified using measurements of existing phantoms and previously obtained measurements of human volunteers. 8 refs

  18. Calibration Monitor for Dark Energy Experiments

    Energy Technology Data Exchange (ETDEWEB)

    Kaiser, M. E.

    2009-11-23

    The goal of this program was to design, build, test, and characterize a flight qualified calibration source and monitor for a Dark Energy related experiment: ACCESS - 'Absolute Color Calibration Experiment for Standard Stars'. This calibration source, the On-board Calibration Monitor (OCM), is a key component of our ACCESS spectrophotometric calibration program. The OCM will be flown as part of the ACCESS sub-orbital rocket payload in addition to monitoring instrument sensitivity on the ground. The objective of the OCM is to minimize systematic errors associated with any potential changes in the ACCESS instrument sensitivity. Importantly, the OCM will be used to monitor instrument sensitivity immediately after astronomical observations while the instrument payload is parachuting to the ground. Through monitoring, we can detect, track, characterize, and thus correct for any changes in instrument senstivity over the proposed 5-year duration of the assembled and calibrated instrument.

  19. Research of Camera Calibration Based on DSP

    Directory of Open Access Journals (Sweden)

    Zheng Zhang

    2013-09-01

    Full Text Available To take advantage of the high-efficiency and stability of DSP in the data processing and the functions of OpenCV library, this study brought forward a scheme that camera calibration in DSP embedded system calibration. An arithmetic of camera calibration based on OpenCV is designed by analyzing the camera model and lens distortion. The transplantation of EMCV to DSP is completed and the arithmetic of camera calibration is migrated and optimized based on the CCS development environment and the DSP/BIOS system. On the premise of realizing calibration function, this arithmetic improves the efficiency of program execution and the precision of calibration and lays the foundation for further research of the visual location based on DSP embedded system.

  20. Calibration of beta dosimeter and personal dosimeter

    International Nuclear Information System (INIS)

    This paper introduces ISO standard ISO 6980 which prepared especially for the calibration of beta dosimeter and personal dosimeter. The standard has three aspects including method of production of reference beta particle radiations, calibration fundamentals related to basic quantities characterizing the radiation field, and calibration of area and personal dosemters and the determination of their response as a function of beta radiation energy and angle of incidence. Here particular emphasis is placed on the determination of basic quantity of tissue absorbed dose at a depth of 0.07 mm in the tissue slab phantom and calibration procedure by mean of the calibration quantity of directional dose equivalent H'(0.07, Ω) and personal dose equivalent Hp (0.07, Ω). Finally, combined standard uncertainty for the determination of absorbed dose rate and component uncertainties of calibration is given as examples. (authors)

  1. Calibration biases in logical reasoning tasks

    OpenAIRE

    Guillermo Macbeth; Alfredo López Alonso; Eugenia Razumiejczyk; Rodrigo Sosa; Carolina Pereyra; Humberto Fernández

    2013-01-01

    The aim of this contribution is to present an experimental study about calibration in deductive reasoning tasks. Calibration is defi ned as the empirical convergence or divergence between the objective and the subjective success. The underconfi dence bias is understood as the dominance of the former over the latter. The hypothesis of this study states that the form of the propositions presented in the experiment is critical for calibration phenomena. Affi rmative and negative propositions are...

  2. Calibration of Nacelle-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  3. On Single-scanline Camera Calibration

    OpenAIRE

    Horaud, Radu; Mohr, Roger; Lorecki, Boguslaw

    1993-01-01

    A method for calibrating single scanline CCD cameras is described. It is shown that the more classical 2D camera calibration techniques are necessary but not sufficient for solving the 1D camera calibration problem. A model for single scanline cameras is proposed, and a two-step procedure for estimating its parameters is provided. It is also shown how the extrinsic camera parameters can be determined geometrically without making explicit the intrinsic camera parameters. The accuracy of the ca...

  4. Camera calibration from surfaces of revolution

    OpenAIRE

    Wong, KYK; Mendonça, PRS; Cipolla, R.

    2003-01-01

    This paper addresses the problem of calibrating a pinhole camera from images of a surface of revolution. Camera calibration is the process of determining the intrinsic or internal parameters (i.e., aspect ratio, focal length, and principal point) of a camera, and it is important for both motion estimation and metric reconstruction of 3D models. In this paper, a novel and simple calibration technique is introduced, which is based on exploiting the symmetry of images of surfaces of revolution. ...

  5. Calibration metrology for fixed irradiation sensors

    International Nuclear Information System (INIS)

    After having recalled the regulatory and technical framework of the calibration of radioprotection measurement instruments, and outlined some technical and operational constraints, the authors report the development of an in situ calibration methodology, i.e. without displacement of the sensor. After the presentation of the calibration chain (from the measurement given by a fixed sensor to the reference value given by a primary standard), they indicate the definition and calculation of the different calibration coefficients allowing the linking up of the different levels, and also the taking of uncertainties into account. They finally report the validation of results

  6. The calibration system for the GERDA experiment

    International Nuclear Information System (INIS)

    The GERDA experiment uses the neutrinoless double beta decay to probe three fundamental questions in neutrino physics - Are they Dirac or Majorana particles? What is their absolute mass? What is the mass hierarchy of the three generations? In my talk I present the calibration system for the Ge semiconductor diodes enriched in Ge-76. The system is used to set the energy scale and calibrate the pulse shapes which will be used to further reject background events. The lowest possible background is crucial for the whole experiment and therefore the calibration system must not interfere with the data acquisition phase while at the same time operate efficiently during the calibration runs.

  7. Sloan Digital Sky Survey Photometric Calibration Revisited

    Energy Technology Data Exchange (ETDEWEB)

    Marriner, John; /Fermilab

    2012-06-29

    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  8. Sloan Digital Sky Survey Photometric Calibration Revisited

    Science.gov (United States)

    Marriner, J.

    2016-05-01

    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  9. The long term stability of lidar calibrations

    DEFF Research Database (Denmark)

    Courtney, Michael; Gayle Nygaard, Nicolai

    Wind lidars are now used extensively for wind resource measurements. One of the requirements for the data to be accepted in support of project financing (so-called ‘banka-bility’) is to demonstrate the long-term stability of lidar cali-brations. Calibration results for six Leosphere WindCube li-dars......-ters pertaining in the different calibration periods. This is supported by sliding-window analyses of one lidar at one location where the same order of variation is observed as between pre-service and post-service calibrations....

  10. Research of Camera Calibration Based on DSP

    OpenAIRE

    Zheng Zhang; Yukun Wan; Lixin Cai

    2013-01-01

    To take advantage of the high-efficiency and stability of DSP in the data processing and the functions of OpenCV library, this study brought forward a scheme that camera calibration in DSP embedded system calibration. An arithmetic of camera calibration based on OpenCV is designed by analyzing the camera model and lens distortion. The transplantation of EMCV to DSP is completed and the arithmetic of camera calibration is migrated and optimized based on the CCS development environment and the ...

  11. Calibration services for medical applications of radiation

    Energy Technology Data Exchange (ETDEWEB)

    DeWerd, L.A.

    1993-12-31

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs.

  12. Calibration services for medical applications of radiation

    International Nuclear Information System (INIS)

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs

  13. Tool calibration system for micromachining system

    Science.gov (United States)

    Miller, Donald M.

    1979-03-06

    A tool calibration system including a tool calibration fixture and a tool height and offset calibration insert for calibrating the position of a tool bit in a micromachining tool system. The tool calibration fixture comprises a yokelike structure having a triangular head, a cavity in the triangular head, and a port which communicates a side of the triangular head with the cavity. Yoke arms integral with the triangular head extend along each side of a tool bar and a tool head of the micromachining tool system. The yoke arms are secured to the tool bar to place the cavity around a tool bit which may be mounted to the end of the tool head. Three linear variable differential transformer's (LVDT) are adjustably mounted in the triangular head along an X axis, a Y axis, and a Z axis. The calibration insert comprises a main base which can be mounted in the tool head of the micromachining tool system in place of a tool holder and a reference projection extending from a front surface of the main base. Reference surfaces of the calibration insert and a reference surface on a tool bar standard length are used to set the three LVDT's of the calibration fixture to the tool reference position. These positions are transferred permanently to a mastering station. The tool calibration fixture is then used to transfer the tool reference position of the mastering station to the tool bit.

  14. Efficiency calibration of low background gamma spectrometer

    International Nuclear Information System (INIS)

    A method of efficiency calibration is described. The authors used standard ores of U, Ra and Th (power form), KCl and Cs-137 sources to do calibration volume-sources which were directly placed on the detector end cap. In such a measuring geometry, it is not necessary to make coincidence-summing correction. The efficiency calibration curve obtained by the method were compared with results measured by Am-241, Cd-109 and Eu-152 calibration sources. The agree in the error of about 5%

  15. FY2008 Calibration Systems Final Report

    Energy Technology Data Exchange (ETDEWEB)

    Cannon, Bret D.; Myers, Tanya L.; Broocks, Bryan T.

    2009-01-01

    The Calibrations project has been exploring alternative technologies for calibration of passive sensors in the infrared (IR) spectral region. In particular, we have investigated using quantum cascade lasers (QCLs) because these devices offer several advantages over conventional blackbodies such as reductions in size and weight while providing a spectral source in the IR with high output power. These devices can provide a rapid, multi-level radiance scheme to fit any nonlinear behavior as well as a spectral calibration that includes the fore-optics, which is currently not available for on-board calibration systems.

  16. KINIK, Absorber Rod Calibration Kinetics

    International Nuclear Information System (INIS)

    1 - Description of program or function: KINIK is an inverse kinetic code that solves the inverse form of the point kinetic equations using the Runge-Kutta method. An optimization procedure is involved to control the time step and to reduce the running time. Up to 24 delayed neutron groups of different types (in case of heavy water as moderator or beryllium as reflector) are considered. KINIK is commonly applied to determine reactivity worths and to calibrate absorber rods. Following a rod drop, neutron flux or power is recorded as a function of time and used as input. 2 - Method of solution: The inverse point kinetic equations are numerically solved for each time step using the Runge-Kutta method. The input data resulting from measurements are first approximated by polynomials of maximum degree 10 using a least-squares approach

  17. del discurso

    Directory of Open Access Journals (Sweden)

    Alexander Mosquera

    2006-01-01

    Full Text Available En varios comunicados quedó reflejada la tensión de la reciente crisis entre Colombia y Venezuela, suscitada a raíz de la captura en diciembre de 2004 de un guerrillero de la FARC en Caracas (Rodrigo Granda. El presente trabajo es un avance de un proyecto de investigación, que persigue como objetivos: 1 describir el discurso implícito en dichos comunicados; 2 identificar los aspectos semánticos y pragmáticos presentes en la confrontación discursiva; y 3 delinear el componente ideológico reflejado en los comunicados. Para el respectivo análisis se siguieron algunos aspectos teóricos y metodológicos de Colle (2000, Bolívar (1998 yMolero de Cabeza (1985. Entre las conclusiones preliminares de este avance se tiene que, con el caso Granda, Colombia recurrió a la vieja política del "fin justifica los medios", a la vez que puso en entredicho el papel de Venezuela en la lucha contra el terrorismo, quizás como respuesta a su política exterior de estrechos vínculos con el gobierno estadounidense

  18. Clustered Calibration: An Improvement to Radio Interferometric Direction Dependent Self-Calibration

    CERN Document Server

    Kazemi, Sanaz; Zaroubi, Saleem

    2013-01-01

    The new generation of radio synthesis arrays, such as LOFAR and SKA, have been designed to surpass existing arrays in terms of sensitivity, angular resolution and frequency coverage. This evolution has led to the development of advanced calibration techniques that ensure the delivery of accurate results at the lowest possible computational cost. However, the performance of such calibration techniques is still limited by the compact, bright sources in the sky, used as calibrators. It is important to have a bright enough source that is well distinguished from the background noise level in order to achieve satisfactory results in calibration. We present "clustered calibration" as a modification to traditional radio interferometric calibration, in order to accommodate faint sources that are almost below the background noise level into the calibration process. The main idea is to employ the information of the bright sources' measured signals as an aid to calibrate fainter sources that are nearby the bright sources...

  19. Design and Calibration of a Cryogenic Blackbody Calibrator at Centimeter Wavelengths

    CERN Document Server

    Kogut, A J; Fixsen, D J; Limon, M; Mirel, P G A; Levin, S; Seiffert, M; Lubin, P M

    2004-01-01

    We describe the design and calibration of an external cryogenic blackbody calibrator used for the first two flights of the Absolute Radiometer for Cosmology, Astrophysics, and Diffuse Emission (ARCADE) instrument. The calibrator consists of a microwave absorber weakly coupled to a superfluid liquid helium bath. Half-wave corrugations viewed 30 deg off axis reduce the return loss below -35 dB. Ruthenium oxide resistive thermometers embedded within the absorber monitor the temperature across the face of the calibrator. The thermal calibration transfers the calibration of a reference thermometer to the flight thermometers using the flight thermometer readout system. Data taken near the superfluid transition in 8 independent calibrations 4 years apart agree within 0.3 mK, providing an independent verification of the thermometer calibration at temperatures near that of the cosmic microwave background.

  20. The physical principles of XRF calibrations

    International Nuclear Information System (INIS)

    Full text: XRF Control and calibration software has come a long way in recent years. Advances in the multiple regression software sophistication and speed of computers have provided an essential resource to the XRF analyst. Over recent years there has been a trend amongst some analysts to develop XRF calibrations based exclusively on the statistical information given by calibration software. Multiple regression statistics are designed for non-correlated data sets but give unpredictable results if there are significant correlations in the standards used. This is typical for calibrations weighted towards certified reference materials (CRM's) which, being natural materials, contain correlated concentrations. Purely statistical methods in calibration development have applicability only over very short concentration ranges and for materials whose composition varies little. Beyond these ranges, the calibration has the potential to be unstable and has been known to produce significant deviations in analysis of unknown samples. The statistical information generated during XRF calibrations can be a very useful tool when used in conjunction with knowledge of the physics behind the correction factors applied. The matrix coefficients represent physical absorption/enhancement effects within the sample and are not arbitrary numbers used to get a good fit to the calibration line. Inappropriate use of matrix factors and overlap factors can produce low RMS values but erroneous results in unknown samples. This talk will contain examples to demonstrate hazards with different calibration strategies and will include coverage of the following topics: physical effects occurring within the sample as a result of X-ray irradiation; use of multiple regression statistics and what role it plays in the calibration; calibration strategies using synthetic and CRM standards; determining appropriate theoretical and semi-empirical matrix corrections and line overlap factors. Copyright (1999

  1. Energy calibration by silicon resonance: Completing system calibration with one reference material

    Energy Technology Data Exchange (ETDEWEB)

    Healy, M.J.F. [Cranfield University, Defence Academy of the United Kingdom, Shrivenham, Swindon, Wiltshire SN6 8LA (United Kingdom)]. E-mail: m.j.f.healy@cranfield.ac.uk

    2006-08-15

    Accelerator energy calibration is seldom a swift process. A rapid calibration method using the proton elastic scattering resonances of silicon was compared to the gamma resonance method. The silicon method proved rapid and convenient given that silicon is also being explored as a means of solid angle calibration, but only moderately accurate. {sup 13}C was identified as a particularly good material for calibration by the superior and established method of gamma resonance.

  2. Tecnolog??as del cuidado del mundo

    OpenAIRE

    Higuera Esp??n, Javier de la

    2011-01-01

    Desde la ??ptica de la ontolog??a de la actualidad, este texto pretende pensar en la posibilidad de unas ???tecnolog??as del mundo??? que, en correlaci??n con las ???tecnolog??as del yo???, sean responsables del cuidado del acontecer singular que define nuestro presente, amenazado por fuertes tendencias homogeneizadoras capaces de cerrar o compactar el mundo. Por el contrario, el mundo s??lo es mundo si ???en ??l puede abrirse un afuera??? (J.-L. Nancy). Para ello analiza la naturaleza de es...

  3. El Banco del Sur

    OpenAIRE

    Espejo, Silvana

    2012-01-01

    En este trabajo abordaremos las características principales del Banco del Sur como una institución financiera sudamericana. Primero presentaremos sus principales características y explicaremos las características de un banco de desarrollo regional. Luego, expondremos el proceso de conformación del mismo haciendo un recorrido por sus principales antecedentes. Por último, cerraremos con una evaluación sobre el estado actual del Banco del Sur. (Párrafo extraído del texto a modo de resumen)...

  4. Radiological standards and calibration laboratory capabilities

    International Nuclear Information System (INIS)

    The Radiological Standards and Calibrations Laboratory, a part of Pacific Northwest Laboratory (PNL), performs calibrations and upholds reference standards necessary to maintain traceability to national radiological standards. The facility supports U.S. Department of Energy (DOE) programs at the Hanford Site, programs sponsored by DOE Headquarters and other federal agencies, radiological protection programs at other DOE sites, and research programs sponsored through the commercial sector. The laboratory is located in the 318 Building of the Hanford Site's 300 Area. The facility contains five major exposure rooms and several laboratories used for exposure work preparation, low-activity instrument calibrations, instrument performance evaluations, instrument maintenance, instrument design and fabrication work, and thermoluminescent and radiochromic dosimetry. The major exposure facilities are a low-scatter room used for neutron and photon exposures, a source well room used for high-volume instrument calibration work, an x-ray facility used for energy response studies, a high-exposure facility used for high-rate photon calibration work, and a beta standards laboratory used for beta energy response studies and beta reference calibrations. Calibrations are routinely performed for personnel dosimeters, health physics instrumentations, photon transfer standards and alpha, beta and gamma field sources used throughout the Hanford Site. This report describes the standards and calibrations laboratory. Photographs that accompany the text appear in the Appendix and are designated Figure A.1 through A.29

  5. An operational analysis of system calibration

    OpenAIRE

    Gaver, Donald Paul; Mutlu, Hasan B.

    1988-01-01

    Mathematical models are proposed for studying the impact of mis-calibration upon operational effectiveness. Methodology for assessing the system effectiveness and an approach for optimizing the effectiveness of a calibration program are examined. The theory and application are discussed, and the results of some specific and convenient models are presented.

  6. Calibration of Avent Wind IRIS SN 01030167

    DEFF Research Database (Denmark)

    Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement un...

  7. Astronomical calibration of the Maastrichtian (Late Cretaceous)

    DEFF Research Database (Denmark)

    Husson, Dorothée; Galbrun, Bruno; Laskar, Jacques;

    2011-01-01

    Recent improvements to astronomical modeling of the Solar System have contributed to important refinements of the Cenozoic time scale through astronomical calibration of sedimentary series. We extend this astronomical calibration into the Cretaceous, on the base of the 405 ka orbital eccentricity...

  8. Microfabricated field calibration assembly for analytical instruments

    Science.gov (United States)

    Robinson, Alex L.; Manginell, Ronald P.; Moorman, Matthew W.; Rodacy, Philip J.; Simonson, Robert J.

    2011-03-29

    A microfabricated field calibration assembly for use in calibrating analytical instruments and sensor systems. The assembly comprises a circuit board comprising one or more resistively heatable microbridge elements, an interface device that enables addressable heating of the microbridge elements, and, in some embodiments, a means for positioning the circuit board within an inlet structure of an analytical instrument or sensor system.

  9. Lidar to lidar calibration phase 2

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the results from phase 2 of a lidar to lidar (L2L) calibration procedure. Phase two of the project included two measurement campaigns conducted at given sites. The purpose was to find out if the lidar-to-lidar calibration procedure can be conducted with similar results at...

  10. Self-calibration and biconvex compressive sensing

    Science.gov (United States)

    Ling, Shuyang; Strohmer, Thomas

    2015-11-01

    The design of high-precision sensing devises becomes ever more difficult and expensive. At the same time, the need for precise calibration of these devices (ranging from tiny sensors to space telescopes) manifests itself as a major roadblock in many scientific and technological endeavors. To achieve optimal performance of advanced high-performance sensors one must carefully calibrate them, which is often difficult or even impossible to do in practice. In this work we bring together three seemingly unrelated concepts, namely self-calibration, compressive sensing, and biconvex optimization. The idea behind self-calibration is to equip a hardware device with a smart algorithm that can compensate automatically for the lack of calibration. We show how several self-calibration problems can be treated efficiently within the framework of biconvex compressive sensing via a new method called SparseLift. More specifically, we consider a linear system of equations {\\boldsymbol{y}}={\\boldsymbol{D}}{\\boldsymbol{A}}{\\boldsymbol{x}}, where both {\\boldsymbol{x}} and the diagonal matrix {\\boldsymbol{D}} (which models the calibration error) are unknown. By ‘lifting’ this biconvex inverse problem we arrive at a convex optimization problem. By exploiting sparsity in the signal model, we derive explicit theoretical guarantees under which both {\\boldsymbol{x}} and {\\boldsymbol{D}} can be recovered exactly, robustly, and numerically efficiently via linear programming. Applications in array calibration and wireless communications are discussed and numerical simulations are presented, confirming and complementing our theoretical analysis.

  11. Calibration biases in logical reasoning tasks

    Directory of Open Access Journals (Sweden)

    Guillermo Macbeth

    2013-08-01

    Full Text Available The aim of this contribution is to present an experimental study about calibration in deductive reasoning tasks. Calibration is defi ned as the empirical convergence or divergence between the objective and the subjective success. The underconfi dence bias is understood as the dominance of the former over the latter. The hypothesis of this study states that the form of the propositions presented in the experiment is critical for calibration phenomena. Affi rmative and negative propositions are distinguished in their cognitive processing. Results suggests that monotonous compound propositions are prone to underconfi dence. An heuristic approach to this phenomenon is proposed. The activation of a monotony heuristic would produce an illusion of simplicity that generates the calibration bias. These evidence is analysed in the context of the metacognitive modeling of calibration phenomena.

  12. DECal: A Spectrophotometric Calibration System For DECam

    CERN Document Server

    Marshall, J L; DePoy, D L; Prochaska, Travis; Allen, Richard; Behm, Tyler W; Martin, Emily C; Veal, Brannon; Villanueva,, Steven; Williams, Patrick; Wise, Jason

    2013-01-01

    DECal is a new calibration system for the CTIO 4 m Blanco telescope. It is currently being installed as part of the Dark Energy Survey and will provide both broadband flat fields and narrowband (about 1 nm bandwidth) spectrophotometric calibration for the new Dark Energy Camera (DECam). Both of these systems share a new Lambertian flat field screen. The broadband flat field system uses LEDs to illuminate each photometric filter. The spectrophotometric calibration system consists of a monochromator-based tunable light source that is projected onto the flat field screen using a custom line-to-spot fiber bundle and an engineered diffuser. Several calibrated photodiodes positioned along the beam monitor the telescope throughput as a function of wavelength. This system will measure the wavelength-dependent instrumental response function of the total telescope+instrument system in the range 300 < lambda < 1100nm. The spectrophotometric calibration will be performed regularly (roughly once per month) to determ...

  13. Calibration concepts for the MUSE integral field

    Science.gov (United States)

    Kelz, Andreas; Bauer, Svend M.; Roth, Martin M.

    2006-06-01

    The phase-A design study of the Calibration Unit (CU) for the Multi-Unit Spectroscopic Explorer (MUSE) is presented. MUSE is an integral-field spectrograph for the 2nd generation of VLT instruments and offers a relative wide integral-field, adaptive-optics assisted spatial resolution, and a wavelength coverage between 465 and 930 nm. MUSE is a project of seven European institutes and is led by the Centre de Recherche Atronomique de Lyon (CRAL). Amongst other work-packages, the Astrophysical Institute Potsdam (AIP) is responsible for the Calibration Unit. The paper describes the calibration requirements, including issues related to spectral, image quality, and geometrical calibration. The opto-mechanical layout of the calibration unit is presented and the use of focal plane masks to evaluate image distortions and PSF degradations is explained.

  14. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  15. Assessing students' metacognitive calibration with knowledge surveys

    Science.gov (United States)

    Lindsey, Beth A.; Nagel, Megan

    2013-01-01

    "Calibration" is an aspect of metacognition that describes how well students assess their own knowledge. One tool that can help to assess student calibration is the knowledge survey (KS). On a KS, students rate their confidence in their ability to answer questions related to course content. A comparison of a student's confidence level with their actual performance on course exams gives an indication of the student's metacognitive calibration. We report on a study that explores students' responses to a KS in introductory physics and chemistry courses serving both STEM and non-STEM populations. In many courses, Delta (the difference between KS-score and final exam score, a measure of calibration) was anti-correlated with final exam performance. No relationship was found between Delta and students' scientific reasoning abilities. We also report preliminary findings on how calibration differs for questions of a quantitative nature vs. those of a more conceptual nature.

  16. The ATLAS Electromagnetic Calorimeter Calibration Workshop

    CERN Multimedia

    Hong Ma; Isabelle Wingerter

    The ATLAS Electromagnetic Calorimeter Calibration Workshop took place at LAPP-Annecy from the 1st to the 3rd of October; 45 people attended the workshop. A detailed program was setup before the workshop. The agenda was organised around very focused presentations where questions were raised to allow arguments to be exchanged and answers to be proposed. The main topics were: Electronics calibration Handling of problematic channels Cluster level corrections for electrons and photons Absolute energy scale Streams for calibration samples Calibration constants processing Learning from commissioning Forty-five people attended the workshop. The workshop was on the whole lively and fruitful. Based on years of experience with test beam analysis and Monte Carlo simulation, and the recent operation of the detector in the commissioning, the methods to calibrate the electromagnetic calorimeter are well known. Some of the procedures are being exercised in the commisssioning, which have demonstrated the c...

  17. Parallel Calibration for Sensor Array Radio Interferometers

    CERN Document Server

    Brossard, Martin; Pesavento, Marius; Boyer, Rémy; Larzabal, Pascal; Wijnholds, Stefan J

    2016-01-01

    In order to meet the theoretically achievable imaging performance, calibration of modern radio interferometers is a mandatory challenge, especially at low frequencies. In this perspective, we propose a novel parallel iterative multi-wavelength calibration algorithm. The proposed algorithm estimates the apparent directions of the calibration sources, the directional and undirectional complex gains of the array elements and their noise powers, with a reasonable computational complexity. Furthermore, the algorithm takes into account the specific variation of the aforementioned parameter values across wavelength. Realistic numerical simulations reveal that the proposed scheme outperforms the mono-wavelength calibration scheme and approaches the derived constrained Cram\\'er-Rao bound even with the presence of non-calibration sources at unknown directions, in a computationally efficient manner.

  18. Systems and methods of eye tracking calibration

    DEFF Research Database (Denmark)

    2014-01-01

    parameters relate to a calibration of a calculation of gaze information of a user of the device, where the gaze information indicates where the user is looking. While the one or more objects are displayed, eye movement information associated with the user is determined, which indicates eye movement of one...... or more eye features associated with at least one eye of the user. The eye movement information is associated with a first object location of the one or more objects. The one or more calibration parameters are calculated based on the first object location being associated with the eye movement information.......Methods and systems to facilitate eye tracking control calibration are provided. One or more objects are displayed on a display of a device, where the one or more objects are associated with a function unrelated to a calculation of one or more calibration parameters. The one or more calibration...

  19. Absolute sensitivity calibration of extreme ultraviolet photoresists

    Energy Technology Data Exchange (ETDEWEB)

    Jones, Juanita; Naulleau, Patrick P.; Gullikson, Eric M.; Aquila, Andrew; George, Simi; Niakoula, Dimitra

    2008-05-16

    One of the major challenges facing the commercialization of extreme ultraviolet (EUV) lithography remains simultaneously achieving resist sensitivity, line-edge roughness, and resolution requirement. Sensitivity is of particular concern owing to its direct impact on source power requirements. Most current EUV exposure tools have been calibrated against a resist standard with the actual calibration of the standard resist dating back to EUV exposures at Sandia National Laboratories in the mid 1990s. Here they report on an independent sensitivity calibration of two baseline resists from the SEMATECH Berkeley MET tool performed at the Advanced Light Source Calibrations and Standards beamline. The results show the baseline resists to be approximately 1.9 times faster than previously thought based on calibration against the long standing resist standard.

  20. Calibration of phase detector using IQ modulator

    International Nuclear Information System (INIS)

    The beam energy of the J-PARC Linac is calculated by TOF (Time-of-Flight) method with the flight distance and beam phases at the two of measurement points. Because the accuracy of the beam energy measurement is directly depending on the errors in the phase measurement system, all 111 beam phase monitors are calibrated annually. Here, we adopted a calibration method using the IQ modulator as a method for carrying out more simply and accurately, calibration of the phase detection circuit is provided to the phase detection system. In the calibration, we have used the trombone circuit for the adjustment of the reference frequency, but it is thought that the procedure using an IQ modulator is more simple and accurate, and it reduces the time for the calibration. We describe the procedure of the phase detection system and the method of energy calculation. In addition, the general descriptions of the IQ modulation specification and its performances are introduced. (author)

  1. Calibration of Sr-90 ophthalmic applicators

    International Nuclear Information System (INIS)

    The purpose of this paper is to alert users of Sr-90 ophthalmic applicators about potential large errors in calibration of these devices. A discrepancy of more than 50% in calibration of Sr-90 ophthalmic applicators between the US National Institute of Standards and Technology (NIST) and one foreign manufacturer (the world's only remaining supplier) has been reported. A single-plane Sr-90 ophthalmic applicator was calibrated by the manufacturer, by NIST, and by the University of Wisconsin. The manufacturer's close rate calibration is confined to a 3-mm-diameter active area, while NIST measures all beta radiation emitted into a 2-PI solid angle. The discrepancy was verified by means of a technique based on that of NIST. Reports of calibrations at NIST of applicators made by several American manufacturers (no longer available) indicate that large discrepancies exist for other manufacturers as well

  2. Guidelines on calibration of neutron measuring devices

    International Nuclear Information System (INIS)

    The International Atomic Energy Agency and the World Health Organization have agreed to establish an IAEA/WHO Network of Secondary Standard Dosimetry Laboratories (SSDLs) in order to improve accuracy in applied radiation dosimetry throughout the world. These SSDLs must be equipped with, and maintain, secondary standard instruments, which have been calibrated against primary standards, and must be nominated by their governments for membership of the network. The majority of the existing SSDLs were established primarily to work with photon radiation (X-rays and gamma rays). Neutron sources are, however, increasingly being applied in industrial processes, research, nuclear power development and radiation biology and medicine. Thus, it is desirable that the SSDLs in countries using neutron sources on a regular basis should also fulfil the minimum requirements to calibrate neutron measuring devices. It is the primary purpose of this handbook to provide guidance on calibration of instruments for radiation protection. A calibration laboratory should also be in a position to calibrate instrumentation being used for the measurement of kerma and absorbed dose and their corresponding rates. This calibration is generally done with photons. In addition, since each neutron field is usually contaminated by photons produced in the source or by scatter in the surrounding media, neutron protection instrumentation has to be tested with respect to its intrinsic photon response. The laboratory will therefore need to possess equipment for photon calibration. This publication deals primarily with methods of applying radioactive neutron sources for calibration of instrumentation, and gives an indication of the space, manpower and facilities needed to fulfil the minimum requirements of a calibration laboratory for neutron work. It is intended to serve as a guide for centres about to start on neutron dosimetry standardization and calibration. 94 refs, 8 figs, 12 tabs

  3. Selection of stars to calibrate Gaia

    Science.gov (United States)

    Carrasco, J. M.; Voss, H.; Jordi, C.; Fabricius, C.; Pancino, E.; Altavilla, G.

    2015-05-01

    Gaia is an all-sky survey satellite, launched by ESA on 19th December 2013, to obtain parallaxes and proper motions to microarcsecond level precision, radial velocities and astrophysical parameters for about one billion objects down to a limiting magnitude of 20. The chosen strategy to perform the photometric calibration is to split the process into two steps, internal and external calibration. The internal calibration will combine all different transits of a given source to a common reference internal system producing a 'mean' Gaia observation. This internal calibration accounts for the differential instrumental effects (in sensitivity, aperture, PSF, etc.). They depend on the colour and type of the source. For this reason, a selection of calibration sources ensuring a good representation of all kind of observed sources is needed. The entire magnitude and colour range of the sources have to be covered by these calibration stars and for all calibration intervals. It is a challenge to obtain a suitable colour distribution for the standards, especially for bright sources and the daily large scale calibration intervals. Once the mean Gaia observations are produced, a final step, the external calibration, transforms them to absolute fluxes and wavelengths. In principle, few calibration sources are needed (about 200 spectrophotometric standard stars, SPSS, are currently being considered). They need to have accurate determinations of their absolute fluxes and their non-variability need to be ensured below 1% precision. For this purpose, a big international observational effort is being done (using telescopes as 2.2m@CAHA, TNG@LaPalma, NTT@LaSilla, LaRuca@SPM, and others). During this observational effort some cases of non-expected variability of the SPSS candidates have been discovered.

  4. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR

    International Nuclear Information System (INIS)

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba

  5. SMAP Radar Processing and Calibration

    Science.gov (United States)

    West, R.; Jaruwatanadilok, S.; Kwoun, O.; Chaubell, M.

    2013-01-01

    The Soil Moisture Active Passive (SMAP) mission is part of the NASA space-based Earth observation program, and consists of an L-band radar and radiometer scheduled for launch into sun synchronous orbit in late 2014. A joint effort of the Jet Propulsion Laboratory (JPL) and the Goddard Space Flight Center (GSFC), the SMAP mission draws heavily on the design and risk reduction heritage of the Hydrosphere State (Hydros) mission [1], [2]. The SMAP science and applications objectives are to: 1) understand processes that link the terrestrial water, energy and carbon cycles, 2) estimate global water and energy fluxes at the land surface, 3) quantify net carbon flux in boreal landscapes, 4) enhance weather and climate forecast skill, and 5) develop improved flood prediction and drought monitoring capability. To meet these science objectives, SMAP ground processing will combine the attributes of the radar and radiometer observations (in terms of their spatial resolution and sensitivity to soil moisture, surface roughness, and vegetation) to estimate soil moisture with 4% volumetric accuracy at a resolution of 10 km, and freeze-thaw state at a resolution of 1-3 km. Model sensitivities translate the soil moisture accuracy to a radar backscatter accuracy of 1 dB (1 sigma) at 3 km resolution and a brightness temperature accuracy of 1.3 K at 40 km resolution. This paper will describe the level 1 radar processing and calibration challenges and the choices made so far for the algorithms and software implementation.

  6. A Cryogenic Infrared Calibration Target

    CERN Document Server

    Wollack, Edward J; Rinehart, Stephan A

    2014-01-01

    A compact cryogenic calibration target is presented that has a peak diffuse reflectance, $R \\le 0.003$, from $800-4,800\\,{\\rm cm}^{-1}$ $(12-2\\,\\mu$m). Upon expanding the spectral range under consideration to $400-10,000\\,{\\rm cm}^{-1}$ $(25-1\\,\\mu$m) the observed performance gracefully degrades to $R \\le 0.02$ at the band edges. In the implementation described, a high-thermal-conductivity metallic substrate is textured with a pyramidal tiling and subsequently coated with a thin lossy dielectric coating that enables high absorption and thermal uniformity across the target. The resulting target assembly is lightweight, has a low-geometric profile, and has survived repeated thermal cycling from room temperature to $\\sim4\\,$K. Basic design considerations, governing equations, and test data for realizing the structure described are provided. The optical properties of selected absorptive materials -- Acktar Fractal Black, Aeroglaze Z306, and Stycast 2850 FT epoxy loaded with stainless steel powder -- are character...

  7. Fabrication and Calibration of FORTIS

    Science.gov (United States)

    Fleming, Brian T.; McCandliss, Stephan R.; Kaiser, Mary Elizabeth; Kruk, Jeffery; Feldman, Paul D.; Kutyrev, Alexander S.; Li, Mary J.; Rapchun, David A.; Lyness, Eric; Moseley, S. H.; Siegmund, Oswald; Vallerga, John; Martin, Adrian

    2011-01-01

    The Johns Hopkins University sounding rocket group is entering the final fabrication phase of the Far-ultraviolet Off Rowland-circle Telescope for Imaging and Spectroscopy (FORTIS); a sounding rocket borne multi-object spectro-telescope designed to provide spectral coverage of 43 separate targets in the 900 - 1800 Angstrom bandpass over a 30' x 30' field-of-view. Using "on-the-fly" target acquisition and spectral multiplexing enabled by a GSFC microshutter array, FORTIS will be capable of observing the brightest regions in the far-UV of nearby low redshift (z approximately 0.002 - 0.02) star forming galaxies to search for Lyman alpha escape, and to measure the local gas-to-dust ratio. A large area (approximately 45 mm x 170 mm) microchannel plate detector built by Sensor Sciences provides an imaging channel for targeting flanked by two redundant spectral outrigger channels. The grating is ruled directly onto the secondary mirror to increase efficiency. In this paper, we discuss the recent progress made in the development and fabrication of FORTIS, as well as the results of early calibration and characterization of our hardware, including mirror/grating measurements, detector performance, and early operational tests of the micro shutter arrays.

  8. HIBP calibration on WEGA stellarator

    Energy Technology Data Exchange (ETDEWEB)

    Podoba, Yuriy; Otte, Matthias; Wagner, Friedrich [Max-Planck-Institut fuer Plasmaphysik, Greifswald (Germany); Zhezhera, Alexander; Chmyga, Alexander; Kozachok, Alexander; Komarov, Alexander; Bondarenko, Ivan; Deshko, Galina; Khrebtov, Sergey; Krupnik, Ludmila [Kharkov Institute of Plasma Physics, Kharkov (Ukraine)

    2009-07-01

    The heavy ion beam probe (HIBP) is an established non-perturbing diagnostic for determining spatial distributions of plasma potential, density, temperature and poloidal magnetic field of magnetically confined plasmas. These quantities can be determined from the change in the ion beam parameters (charge, intensity and trajectory) passing through a plasma volume due to collisions with electrons and interactions with the confining magnetic field. The WEGA HIBP operates with a Na{sup +} with an energy of 39.5 keV and beam current 35 {mu}A. Conventionally the coordinate mapping of the HIBP is provided by ray tracing calculations of the ion beam in the magnetic field. However, it is very difficult to include all physical effects in the model, thus the result of the calculations may significantly differ from the real probing position. In order to improve the mapping precision an additional measurements of the beam position have been provided using a beam detector array inside the vacuum vessel. This allows to compare the measured and calculated ion beam position in order to find out the reasons for the coordinate mismatch and include adjustments in the calculation code. Results of this calibration are presented in this work.

  9. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  10. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities

    International Nuclear Information System (INIS)

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  11. Accesibilidad del sitio web del CONAPRED

    OpenAIRE

    Luján Mora, Sergio

    2012-01-01

    Accesibilidad del sitio web del CONAPRED (Consejo Nacional para Prevenir la Discriminación) de México. Más información: - http://accesibilidadweb.dlsi.ua.es - http://desarrolloweb.dlsi.ua.es Autor: Sergio Luján Mora, profesor de la Universidad de Alicante (http://www.ua.es).

  12. Geologia del Baix Ter

    OpenAIRE

    Mas-Pla, Josep

    1986-01-01

    Article que analitza el valor i la singularitat de les zones humides del Baix Ter, tenint en compte la descripció de les característiques geològiques de la Plana del Ter i dels diferents ambients morfodinàmics que s' hi troben

  13. Il dono del racconto

    DEFF Research Database (Denmark)

    Polito, Paola

    1998-01-01

    L'articolo indaga a vari livelli strutturali del romanzo "Il coraggio del pettirosso" (rete metaforica, dialogismo diegetico, dialogismo come fenomeno di voce, strategie discorsive messe in atto dalla voce narrante, stile colloquiale) la complessa realizzazione testuale del genere narrativo /racc...

  14. Preliminary evaluation of a Neutron Calibration Laboratory

    International Nuclear Information System (INIS)

    In the past few years, Brazil and several other countries in Latin America have experimented a great demand for the calibration of neutron detectors, mainly due to the increase in oil prospection and extraction. The only laboratory for calibration of neutron detectors in Brazil is localized at the Institute for Radioprotection and Dosimetry (IRD/CNEN), Rio de Janeiro, which is part of the IAEA SSDL network. This laboratory is the national standard laboratory in Brazil. With the increase in the demand for the calibration of neutron detectors, there is a need for another calibration services. In this context, the Calibration Laboratory of IPEN/CNEN, Sao Paulo, which already offers calibration services of radiation detectors with standard X, gamma, beta and alpha beams, has recently projected a new calibration laboratory for neutron detectors. In this work, the ambient equivalent dose rate (H⁎(10)) was evaluated in several positions inside and around this laboratory, using Monte Carlo simulation (MCNP5 code), in order to verify the adequateness of the shielding. The obtained results showed that the shielding is effective, and that this is a low-cost methodology to improve the safety of the workers and evaluate the total staff workload. (author)

  15. Increased Automation in Stereo Camera Calibration Techniques

    Directory of Open Access Journals (Sweden)

    Brandi House

    2006-08-01

    Full Text Available Robotic vision has become a very popular field in recent years due to the numerous promising applications it may enhance. However, errors within the cameras and in their perception of their environment can cause applications in robotics to fail. To help correct these internal and external imperfections, stereo camera calibrations are performed. There are currently many accurate methods of camera calibration available; however, most or all of them are time consuming and labor intensive. This research seeks to automate the most labor intensive aspects of a popular calibration technique developed by Jean-Yves Bouguet. His process requires manual selection of the extreme corners of a checkerboard pattern. The modified process uses embedded LEDs in the checkerboard pattern to act as active fiducials. Images are captured of the checkerboard with the LEDs on and off in rapid succession. The difference of the two images automatically highlights the location of the four extreme corners, and these corner locations take the place of the manual selections. With this modification to the calibration routine, upwards of eighty mouse clicks are eliminated per stereo calibration. Preliminary test results indicate that accuracy is not substantially affected by the modified procedure. Improved automation to camera calibration procedures may finally penetrate the barriers to the use of calibration in practice.

  16. Calibration of Ground-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...... measurements are given for information only....

  17. Optimal, Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard

    Reliability based code calibration is considered in this paper. It is described how the results of FORM based reliability analysis may be related to the partial safety factors and characteristic values. The code calibration problem is presented in a decision theoretical form and it is discussed how...... acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for the ultimate and the serviceability limit states. Finally the paper describes a procedure for the practical implementation of...... reliability based code calibration of LRFD based design codes....

  18. Single Camera Calibration in 3D Vision

    OpenAIRE

    Caius SULIMAN; Puiu, Dan; Moldoveanu, Florin

    2009-01-01

    Camera calibration is a necessary step in 3D vision in order to extract metric information from 2D images. A camera is considered to be calibrated when the parameters of the camera are known (i.e. principal distance, lens distorsion, focal length etc.). In this paper we deal with a single camera calibration method and with the help of this method we try to find the intrinsic and extrinsic camera parameters. The method was implemented with succes in the programming and simulation environment M...

  19. Multi-Camera Calibration Using a Globe

    OpenAIRE

    Shen, Rui; Cheng, Irene; Basu, Anup

    2008-01-01

    The need for calibration of multiple cameras working together in a network, or for the acquisition of free viewpoint video for 3D TV, is becoming increasingly important in recent years. In this paper we present a novel approach for calibrating multiple cameras using an ordinary globe that is usually available in every household. This method makes it possible to reduce multi-camera calibration to a level that is attainable by non-technical users. Our technique requires only one view of the glo...

  20. Calibration for 3D Structured Light Measurement

    Institute of Scientific and Technical Information of China (English)

    2007-01-01

    A calibration procedure was developed for three-dimensional(3D) binocular structured light measurement systems. In virtue of a specially designed pattern, matching points in stereo images are extracted. And then sufficient 3D space points are obtained through pairs of images with the intrinsic and extrinsic parameters of each camera estimated prior and consequently some lights are calibrated by means of multi point fitting. Finally, a mathematical model is applied to interpolate and approximate all dynamic scanning lights based on geometry. The process of calibration method is successfully used in the binocular 3D measurement system based on structured lights and the 3D reconstruction results are satisfying.

  1. Calibration of personal dosimeters: Quantities and terminology

    International Nuclear Information System (INIS)

    The numerical results obtained in the interpretation of individual monitoring of external radiation depend not only on the accurate calibration of the radiation measurement instruments involved, but also on the definition of the quantities in term of which these instruments are calibrated The absence of uniformity in terminology not only makes it difficult to understand properly the scientific and technical literature but can also lead to incorrect interpretation of particular concepts and recommendations. In this paper, brief consideration is given to definition of radiation quantities and terminology used in calibration procedures. (author)

  2. Calibration boards for the ATLAS LAr calorimeters

    International Nuclear Information System (INIS)

    In order to calibrate the ATLAS Liquid Argon (LAr) calorimeters to an accuracy better than 1%, over 16 bit dynamic range, chips have been designed in DMILL technology. The design and performance of a 16 bit DAC, a static low offset operational amplifier and a digital chip to control the calibration boards are presented. A 8 channels board using these chips has also been realised and carefully measured as this module will be replicated 16 times to design the final 128 channels calibration board. (authors)

  3. Global analysis of the phase calibration operation

    Science.gov (United States)

    Lannes, André

    2005-04-01

    A global approach to phase calibration is presented. The corresponding theoretical framework calls on elementary concepts of algebraic graph theory (spanning tree of maximal weight, cycles) and algebraic number theory (lattice, nearest lattice point). The traditional approach can thereby be better understood. In radio imaging and in optical interferometry, the self-calibration procedures must often be conducted with much care. The analysis presented should then help in finding a better compromise between the coverage of the calibration graph (which must be as complete as possible) and the quality of the solution (which must of course be reliable).

  4. Design of a neutron source for calibration

    International Nuclear Information System (INIS)

    The neutron spectra produced by an isotopic neutron source located at the center of moderating media were calculated using Monte Carlo method in the aim to design a neutron source for calibration purposes. To improve the evaluation of the dosimetric quantities, is recommended to calibrate the radiation protection devices with calibrated neutron sources whose neutron spectra being similar to those met in practice. Here, a 239Pu-Be neutron source was inserted in H2O, D2O and polyethylene cylindrical moderators in order to produce neutron spectra that resembles spectra found in workplaces

  5. The ENEA calibration service for ionising radiations; Il centro di taratura per le radiazioni ionizzanti di Bologna. Parte 1: Fotoni

    Energy Technology Data Exchange (ETDEWEB)

    Monteventi, F.; Sermenghi, I. [ENEA, Centro Ricerche Ezio Clementel, Bologna (Italy). Dipt. Ambiente

    1999-07-01

    The report describes all the facilities available at the the service of the ENEA Calibration Service for Ionising Radiations at Bologna (Italy). It gives a detailed description of all equipment qualified for photon fields metrology including the secondary standards and the calibration procedures performed for radiation monitoring devices and dosemeters. [Italian] Il presente lavoro descrive i servizi e le attivita' del Centro di Taratura dell'ENEA di Bologna, in particolare delle attrezzature qualificate per la metrologia fotonica, dei campioni di misura e delle procedure adottate per la taratura degli strumenti e dei dosimetri.

  6. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    Directory of Open Access Journals (Sweden)

    W. Jessen

    2015-10-01

    Full Text Available Resource assessment for Concentrated Solar Power (CSP needs accurate Direct Normal Irradiance (DNI measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  7. Calibration of routine dosimeters in radiation processing: Validation procedure for in-plant calibration

    Directory of Open Access Journals (Sweden)

    Šećerov Bojana Lj.

    2011-01-01

    Full Text Available The essential prerequisite of radiation dosimetry is to provide quality assurance and documentation that the irradiation procedure has been carried out according to the specification requirement of correct calibration of the chosen dosimetry system. At the Radiation Plant of the Vinča Institute of Nuclear Sciences we compared two recommended protocols of irradiation procedures in the calibration of dosimetry systems in radiation processing: (1 by irradiation of routine dosimeters (ethanol-chlorobenzene - ECB at the calibration laboratory and (2, by in-plant calibration with alanine transfer - dosimeters. The critical point for in-plant calibration is irradiation geometry, so we carefully positioned the phantom carrying both dosimeters in order to minimize dose gradients across the sample. The analysis of results obtained showed that the difference among determined absorbed doses for the construction of calibration curves between these two methods, (alanine vs. ECB, is less than 1%. The difference in combined standard uncertainty for each calibration procedure is 0.1%. These results demonstrate that our in-plant calibration is as good as calibration by irradiation at the calibration laboratory and validates our placement of the irradiation phantom during irradiation.

  8. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    Science.gov (United States)

    Jessen, W.; Wilbert, S.; Nouri, B.; Geuder, N.; Fritz, H.

    2015-10-01

    Resource assessment for Concentrated Solar Power (CSP) needs accurate Direct Normal Irradiance (DNI) measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs) with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA) in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May) show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  9. A taxonomy of camera calibration and video projection correction methods

    OpenAIRE

    Radhwan Ben Madhkour; Matei Mancas; Thierry Dutoit

    2015-01-01

    This paper provides a classification of calibration methods for cameras and projectors. From basic homography to complex geometric calibration methods, this paper aims at simplifying the choice of the methods to perform a calibration regarding the complexity of the setup. The classical camera calibration methods are presented. A comparison gives the pros and cons for each method. For the projector calibration, the homography, the structured light methods and the geometric calibration are pres...

  10. Bartolome Island, Galapagos Stable Oxygen Calibration Data

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Galapagos Coral Stable Oxygen Calibration Data. Sites: Bartolome Island: 0 deg, 17'S, 90 deg 33' W. Champion Island: 1 deg, 15'S, 90 deg, 05' W. Urvina Bay (Isabela...

  11. Champion Island, Galapagos Stable Oxygen Calibration Data

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Galapagos Coral Stable Oxygen Calibration Data. Sites: Bartolome Island: 0 deg, 17 min S, 90 deg 33 min W. Champion Island: 1 deg, 15 min S, 90 deg, 05 min W....

  12. Calibration of pressure gauge for Cherenkov detector

    CERN Document Server

    Saponjic, Nevena

    2013-01-01

    Solartron/Hamilton pressure gauges are used to monitor the gas pressure in the particle beam detectors installed in the experimental areas. Here is description of the test bench for the calibration of these gauges in Labview.

  13. Calibration of a photomultiplier array spectrometer

    Science.gov (United States)

    Bailey, Steven A.; Wright, C. Wayne; Piazza, Charles R.

    1989-01-01

    A systematic approach to the calibration of a photomultiplier array spectrometer is presented. Through this approach, incident light radiance derivation is made by recognizing and tracing gain characteristics for each photomultiplier tube.

  14. Calibration device for ultrasonic inspection equipment

    International Nuclear Information System (INIS)

    The inspection equipment is introduced into the reactor vessel and deposited there. The real inspection of the vessel is performed by means of a manipulator arm carrying an ultrasonic transducer field and being desplaceable along nine axes of motion for the inspection of welds. In order to be sure of adequate testing the origin of the field must exactly be known. This is achieved by means of a calibration unit by which the exact zero position in the vessel is fixed, the position of the ultrasonic transducer is calibrated and its angular position can be determined. It has got a spherical calibration body and at least one inlined reflection surface (tuncated cone surface) that can take a fixed position with respect to the ultrasonic transducer. The calibration body and/or the reflection surface are moved with respect to the ultrasonic transducer until the maximum amplitude of the reflected signal is obtained. (DG)

  15. A transmission calibration method for superconducting resonators

    CERN Document Server

    Cataldo, Giuseppe; Barrentine, Emily M; Brown, Ari D; Moseley, Samuel H; U-Yen, Kongpop

    2014-01-01

    A method is proposed and experimentally explored for \\textit{in-situ} calibration of complex transmission data for superconducting microwave resonators. This cryogenic calibration method accounts for the instrumental transmission response between the vector network analyzer reference plane and the device calibration plane. Once calibrated, the observed resonator response was modeled in detail by two approaches. The first, a phenomenological model based on physically realizable rational functions, enables the extraction of multiple resonance frequencies and widths for coupled resonators without explicit specification of the circuit network. In the second, an ABCD-matrix representation for the distributed transmission line circuit is used to model the observed response from the characteristic impedance and propagation constant. When used in conjunction with electromagnetic simulations, the kinetic inductance fraction can be determined with this method with an accuracy of 2%. Datasets for superconducting microst...

  16. Entry tank calibration in TOR pilot plant

    International Nuclear Information System (INIS)

    The objective of this communication is the description of calibration measurements used for determining the uranium and plutonium mass entry in the fast neutron fuel reprocessing pilot plant (TOR) of Marcoule

  17. Construction of calibration curve for accountancy tank

    International Nuclear Information System (INIS)

    Tanks are equipped in a reprocessing plant for accounting solution of nuclear material. The careful measurement of volume in tanks is very important to implement rigorous accounting of nuclear material. The calibration curve relating the volume and level of solution needs to be constructed, where the level is determined by differential pressure of dip tubes. Several calibration curves are usually employed, but it's not explicitly decided how many segment are used, where to select segment, or what should be the degree of polynomial curve. These parameters, i.e., segment and degree of polynomial curve are mutually interrelated to give the better performance of calibration curve. Here we present the construction technique of giving optimum calibration curves and their characteristics. (author)

  18. The 2007 ESO Instrument Calibration Workshop

    CERN Document Server

    Kaufer, Andreas; ESO Workshop

    2008-01-01

    The 2007 ESO Instrument Calibration workshop brought together more than 120 participants with the objective to a) foster the sharing of information, experience and techniques between observers, instrument developers and instrument operation teams, b) review the actual precision and limitations of the applied instrument calibration plans, and c) collect the current and future requirements by the ESO users. These present proceedings include the majority of the workshop’s contributions and document the status quo of instrument calibration at ESO in large detail. Topics covered are: Optical Spectro-Imagers, Optical Multi-Object Spectrographs, NIR and MIR Spectro-Imagers, High-Resolution Spectrographs, Integral Field Spectrographs, Adaptive Optics Instruments, Polarimetric Instruments, Wide Field Imagers, Interferometric Instruments as well as other crucial aspects such as data flow, quality control, data reduction software and atmospheric effects. It was stated in the workshop that "calibration is a life-long l...

  19. Laboratory calibration and characterization of video cameras

    Science.gov (United States)

    Burner, A. W.; Snow, W. L.; Shortis, M. R.; Goad, W. K.

    1990-01-01

    Some techniques for laboratory calibration and characterization of video cameras used with frame grabber boards are presented. A laser-illuminated displaced reticle technique (with camera lens removed) is used to determine the camera/grabber effective horizontal and vertical pixel spacing as well as the angle of nonperpendicularity of the axes. The principal point of autocollimation and point of symmetry are found by illuminating the camera with an unexpanded laser beam, either aligned with the sensor or lens. Lens distortion and the principal distance are determined from images of a calibration plate suitably aligned with the camera. Calibration and characterization results for several video cameras are presented. Differences between these laboratory techniques and test range and plumb line calibration are noted.

  20. Non-iterative method for camera calibration.

    Science.gov (United States)

    Hong, Yuzhen; Ren, Guoqiang; Liu, Enhai

    2015-09-01

    This paper presents a new and effective technique to calibrate a camera without nonlinear iteration optimization. To this end, the centre-of-distortion is accurately estimated firstly. Based on the radial distortion division model, point correspondences between model plane and its image were used to compute the homography and distortion coefficients afterwards. Once the homographies of calibration images are obtained, the camera intrinsic parameters are solved analytically. All the solution techniques applied in this calibration process are non-iterative that do not need any initial guess, with no risk of local minima. Moreover, estimation of the distortion coefficients and intrinsic parameters could be successfully decoupled, yielding the more stable and reliable result. Both simulative and real experiments have been carried out to show that the proposed method is reliable and effective. Without nonlinear iteration optimization, the proposed method is computationally efficient and can be applied to real-time online calibration. PMID:26368490

  1. The twisted cubic and camera calibration

    OpenAIRE

    Buchanan, Thomas

    1988-01-01

    We state a uniqueness theorem for camera calibration in terms of the twisted cubic. The theorem assumes the general linear model and is essentially a reformulation of Seydewitz's star generation theorem.

  2. The PREMOS/PICARD instrument calibration

    Science.gov (United States)

    Schmutz, Werner; Fehlmann, André; Hülsen, Gregor; Meindl, Peter; Winkler, Rainer; Thuillier, Gérard; Blattner, Peter; Buisson, François; Egorova, Tatiana; Finsterle, Wolfgang; Fox, Nigel; Gröbner, Julian; Hochedez, Jean-François; Koller, Silvio; Meftah, Mustapha; Meisonnier, Mireille; Nyeki, Stephan; Pfiffner, Daniel; Roth, Hansjörg; Rozanov, Eugene; Spescha, Marcel; Wehrli, Christoph; Werner, Lutz; Wyss, Jules U.

    2009-08-01

    PREMOS is a space experiment scheduled to fly on the French solar mission PICARD. The experiment comprises filter radiometers and absolute radiometers to measure the spectral and total solar irradiance. The aim of PREMOS is to contribute to the long term monitoring of the total solar irradiance, to use irradiance observations for 'nowcasting' the state of the terrestrial middle atmosphere and to provide long term sensitivity calibration for the solar imaging instrument SODISM on PICARD. In this paper we describe the calibration of the instruments. The filter radiometer channels in the visible and near IR were characterized at PMOD/WRC and the UV channels were calibrated at PTB Berlin. The absolute radiometers were compared with the World Radiometric Reference at PMOD/WRC and a power calibration relative to a primary cryogenic radiometer standard was performed in vacuum and air at NPL.

  3. Strain Gauges Mounted To Retain Calibration

    Science.gov (United States)

    Butler, Barry L.

    1993-01-01

    Silicon-based semiconductor strain gauges mounted in such way they retain original calibration for several years instead of few months. Improvement effected by bonding gauges to ceramic substrates with glasses instead of epoxies as adhesives.

  4. Calibration of sources for alpha spectroscopy systems

    International Nuclear Information System (INIS)

    This paper describes the calibration methodology for measuring the total alpha activity of plane and thin sources with the Alpha Spectrometer for Silicon Detector in the Nuclear Measures and Dosimetry laboratory at IEAv/CTA. (author)

  5. Calibration system of fiber bragg gratings measurement

    OpenAIRE

    Jelínek, M. (Martin)

    2015-01-01

    This work deals with design of the calibration part of the nuclear power plant containment shape monitoring system. The design and implementation system of the optical filter thermal compensation is described. The control system was implemented in LabVIEW.

  6. Computer Vision Assisted Virtual Reality Calibration

    Science.gov (United States)

    Kim, W.

    1999-01-01

    A computer vision assisted semi-automatic virtual reality (VR) calibration technology has been developed that can accurately match a virtual environment of graphically simulated three-dimensional (3-D) models to the video images of the real task environment.

  7. Low Power, Self Calibrated Vector Magnetometer Project

    Data.gov (United States)

    National Aeronautics and Space Administration — This Phase I SBIR project investigates a novel approach to vector magnetometry based on high precision measurements of the total magnetic field. The calibration is...

  8. HPS instrument calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Masse, F.X; Eisenhower, E.H.; Swinth, K.L.

    1993-12-31

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory.

  9. Calibration of the JEM-EUSO detector

    Directory of Open Access Journals (Sweden)

    Gorodetzky P.

    2013-06-01

    Full Text Available In order to unveil the mystery of ultra high energy cosmic rays (UHECRs, JEM-EUSO (Extreme Universe Space Observatory on-board Japan Experiment Module will observe extensive air showers induced by UHECRs from the International Space Station orbit with a huge acceptance. Calibration of the JEM-EUSO instrument, which consists of Fresnel optics and a focal surface detector with 5000 photomultipliers, is very important to discuss the origin of UHECRs precisely with the observed results. In this paper, the calibration before launch and on-orbit is described. The calibration before flight will be performed as precisely as possible with integrating spheres. In the orbit, the relative change of the performance will be checked regularly with on-board and on-ground light sources. The absolute calibration of photon detection efficiency may be performed with the moon, which is a stable light source in the nature.

  10. Low radioactivity spectral gamma calibration facility

    International Nuclear Information System (INIS)

    A low radioactivity calibration facility has been constructed at the Nevada Test Site (NTS). This facility has four calibration models of natural stone that are 3 ft in diameter and 6 ft long, with a 12 in. cored borehole in the center of each model and a lead-shielded run pipe below each model. These models have been analyzed by laboratory natural gamma ray spectroscopy (NGRS) and neutron activation analysis (NAA) for their K, U, and Th content. Also, 42 other elements were analyzed in the NAA. The 222Rn emanation data were collected. Calibrating the spectral gamma tool in this low radioactivity calibration facility allows the spectral gamma log to accurately aid in the recognition and mapping of subsurface stratigraphic units and alteration features associated with unusual concentrations of these radioactive elements, such as clay-rich zones

  11. Absolute calibration technique for spontaneous fission sources

    International Nuclear Information System (INIS)

    An absolute calibration technique for a spontaneously fissioning nuclide (which involves no arbitrary parameters) allows unique determination of the detector efficiency for that nuclide, hence of the fission source strength

  12. HPS instrument calibration laboratory accreditation program

    International Nuclear Information System (INIS)

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory

  13. Commissioning the CMS alignment and calibration framework

    International Nuclear Information System (INIS)

    The CMS experiment has developed a powerful framework to ensure the precise and prompt alignment and calibration of its components, which is a major prerequisite to achieve the optimal performance for physics analysis. The prompt alignment and calibration strategy harnesses computing resources both at the Tier-0 site and the CERN Analysis Facility (CAF) to ensure fast turnaround for updating the corresponding database payloads. An essential element is the creation of dedicated data streams concentrating the specific event information required by the various alignment and calibration workflows. The resulting low latency is required for feeding the resulting constants into the prompt reconstruction process, which is essential for achieving swift physics analysis of the LHC data. This report discusses the implementation and the computational aspects of the alignment and calibration framework. Recent commissioning campaigns with cosmic muons, beam halo and simulated data have been used to gain detailed experience with this framework, and results of this validation are reported.

  14. Absolute calibration of JET ELE system

    International Nuclear Information System (INIS)

    The first Michelson channel of the JET ECE system has been calibrated absolutely using a new high temperature source. The estimated uncertainties are of order +- 20% in the absolute spectral response and +- 10% in the relative spectral shape

  15. Regional Location Calibration in Asia

    Science.gov (United States)

    Steck, L. K.; Hartse, H.; Aprea, C.; Franks, J.; Velasco, A.; Randall, G.; Bradley, C.; Begnaud, M.; Aguilar-Chang, J.

    2002-12-01

    This paper presents a spectrum of issues and efforts involved in improving seismic location performance worldwide. Our efforts are largely designed around providing validated, rigorously calibrated travel times, azimuths, and slownesses along with accurate error estimates. To do so entails a significant effort that includes data mining, data integration, database management, developing optimal 1-, 2-, and 3-D Earth models, using the Earth models to predict wave propagation, developing corrections and errors for travel times, azimuths, and slownesses, and validation of all products. Results presented here will focus on Asia. For the region around station MAKZ in north-central Asia we have looked at several tens of published 1-D velocity models. For each model, travel time calculations were performed, predictions for P and S arrivals were established, and the predicted times were compared to the observed. We will present best-fit models for tectonic provinces out to regional distances from MAKZ. Previous work has shown that Non-stationary Modified Bayesian Kriging of travel time residuals successfully improves regional seismic event location, and this method is being extended to calculate corrections for azimuth and slowness. The ability to krig over 3-D Earth models is also being implemented. In order to produce the most useful corrections, we require accurate ground truth. For this we are continuing efforts to create a location database consisting of the best available seismic event locations and the most accurate and precise travel times. Building this database relies on participation from universities, other NNSA laboratories, and contacts in private industry. Through the kriging procedure we are able to stabilize location algorithms, but the ultimate usefulness of the corrections themselves is directly related to the quality of the ground truth from which the corrections are derived. Indeed, epicentral mislocations from EvLoc using travel time correction

  16. 40 CFR 1065.315 - Pressure, temperature, and dewpoint calibration.

    Science.gov (United States)

    2010-07-01

    ...) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Calibrations and Verifications Measurement of Engine... temperature-equilibrated and temperature-monitored calibration salt solutions in containers that...

  17. Automatic colorimetric calibration of human wounds

    Directory of Open Access Journals (Sweden)

    Meert Theo

    2010-03-01

    Full Text Available Abstract Background Recently, digital photography in medicine is considered an acceptable tool in many clinical domains, e.g. wound care. Although ever higher resolutions are available, reproducibility is still poor and visual comparison of images remains difficult. This is even more the case for measurements performed on such images (colour, area, etc.. This problem is often neglected and images are freely compared and exchanged without further thought. Methods The first experiment checked whether camera settings or lighting conditions could negatively affect the quality of colorimetric calibration. Digital images plus a calibration chart were exposed to a variety of conditions. Precision and accuracy of colours after calibration were quantitatively assessed with a probability distribution for perceptual colour differences (dE_ab. The second experiment was designed to assess the impact of the automatic calibration procedure (i.e. chart detection on real-world measurements. 40 Different images of real wounds were acquired and a region of interest was selected in each image. 3 Rotated versions of each image were automatically calibrated and colour differences were calculated. Results 1st Experiment: Colour differences between the measurements and real spectrophotometric measurements reveal median dE_ab values respectively 6.40 for the proper patches of calibrated normal images and 17.75 for uncalibrated images demonstrating an important improvement in accuracy after calibration. The reproducibility, visualized by the probability distribution of the dE_ab errors between 2 measurements of the patches of the images has a median of 3.43 dE* for all calibrated images, 23.26 dE_ab for all uncalibrated images. If we restrict ourselves to the proper patches of normal calibrated images the median is only 2.58 dE_ab! Wilcoxon sum-rank testing (p Conclusion The investigators proposed an automatic colour calibration algorithm that ensures reproducible colour

  18. Biodiversidad del margen continental del Caribe colombiano.

    OpenAIRE

    2010-01-01

    Incluye valiosa información producto de varios años de investigación marina de las aguas profundas, la plataforma continental y el talud superior del Caribe colombiano, llevada a cabo por un grupo multidisciplinario de investigadores en áreas de taxonomía, geología, ecología, cartografía y conservación del Invemar. Dentro de los resultados más destacados se encuentra el haber colectado e identificado más de 1600 especies, de las cuales alrededor del 35% se constituyeron en primeros registr...

  19. Conception of CTMSP ionizing radiation calibration laboratory

    International Nuclear Information System (INIS)

    The present paper describes the implantation process of an ionizing radiation calibration laboratory in a preexistent installation in CTMSP (bunker) approved by CNEN to operate with gamma-ray for non destructive testing. This laboratory will extend and improve the current metrological capacity for the attendance to the increasing demand for services of calibration of ionizing radiation measuring instruments. Statutory and regulatory requirements for the licensing of the installation are presented and deeply reviewed. (author)

  20. First results from the MINOS calibration detector

    CERN Document Server

    Vahle, P; Alner, J; Anderson, B; Attree, D; Barker, M; Belias, A; Crone, G; Durkin, T J; Felt, N; Falk, E; Harris, P; Jenner, L; Kordosky, M; Lang, K; Lebedev, A; Lee, R; Longley, N P; Marshak, M; Miyagawa, P; Michael, D G; Morse, R; Musser, J; Nichol, R; Nicholls, T; Oliver, J; Pearce, G; Petyt, D; Proga, M; Rebel, B; Saakyan, R; Smith, C; Sullivan, P; Thomas, J; Weber, A; Wojcicki, S G

    2002-01-01

    The MINOS calibration detector (CalDet) is a small version of the MINOS Near and Far neutrino detectors. A program of exposure to beams of muons, electrons, pions and protons at the CERN PS will provide calibration of the calorimetric and topological response of the Near and Far detectors. In this talk, we briefly discuss the goals and design of the CalDet and present first results from the initial beam exposure. (3 refs).

  1. Camera calibration from road lane markings

    OpenAIRE

    Fung, GSK; Yung, NHC; Pang, GKH

    2003-01-01

    Three-dimensional computer vision techniques have been actively studied for the purpose of visual traffic surveillance. To determine the 3-D environment, camera calibration is a crucial step to resolve the relationship between the 3-D world coordinates and their corresponding image coordinates. A novel camera calibration using the geometry properties of road lane markings is proposed. A set of equations that computes the camera parameters from the image coordinates of the road lane markings a...

  2. Adaptive visual servoing by simultaneous camera calibration

    OpenAIRE

    Pomares, J.; Chaumette, François; Torres, F.

    2007-01-01

    Calibration techniques allow the estimation of the intrinsic parameters of a camera. This paper describes an adaptive visual servoing scheme which employs the visual data measured during the task to determine the camera intrinsic parameters. This approach is based on the virtual visual servoing approach. However, in order to increase the robustness of the calibration several aspects have been introduced in this approach with respect to the previous developed virtual vi...

  3. Compact Optical Technique for Streak Camera Calibration

    International Nuclear Information System (INIS)

    The National Ignition Facility is under construction at the Lawrence Livermore National Laboratory for the U.S. Department of Energy Stockpile Stewardship Program. Optical streak cameras are an integral part of the experimental diagnostics instrumentation. To accurately reduce data from the streak cameras a temporal calibration is required. This article describes a technique for generating trains of precisely timed short-duration optical pulses that are suitable for temporal calibrations

  4. Projector Calibration Using a Markerless Plane

    OpenAIRE

    Draréni, Jamil; Roy, Sébastien; Sturm, Peter

    2009-01-01

    In this paper we address the problem of geometric video projector calibration using a markerless planar surface (wall) and a partially calibrated camera. Instead of using control points to infer the camera-wall orientation, we find such relation by efficiently sampling the hemisphere of possible orientations. This process is so fast that even the focal of the camera can be estimated during the sampling process. Hence, physical grids and full knowledge of camera parameters are no longer necess...

  5. Identifiable parameters for parallel robots kinematic calibration

    OpenAIRE

    Besnard, Sébastian; Khalil, Wisama

    2001-01-01

    This paper presents a numerical method for the determination of the identifiable parameters of parallel robots. The special case of Stewart-Gough 6 degrees-offreedom parallel robots is studied for classical and self calibration methods, but this method can be generalized to any kind of parallel robot. The method is based on QR decomposition of the observation matrix of the calibration system. Numerical relations between the identifiable and non identifiable parameters can be obtained.

  6. pH sensor calibration procedure

    OpenAIRE

    Artero Delgado, Carola; Nogueras Cervera, Marc; Manuel Lázaro, Antonio; Prat Tasias, Jordi; Prat Farran, Joana d'Arc

    2013-01-01

    This paper describes the calibration of pH sensor located at the OBSEA marine Observatory. This instrument is based on an industrial pH electrode that is connected to a CTD instrument (Conductivity, Temperature, and Depth ). The calibration of the pH sensor has been done using a high precision spectrophotometer pH meter from Institute of Marine Sciences (ICM), and in this way it has been obtained a numerical function for the p H sensor propor...

  7. Hydraulic Calibrator for Strain-Gauge Balances

    Science.gov (United States)

    Skelly, Kenneth; Ballard, John

    1987-01-01

    Instrument for calibrating strain-gauge balances uses hydraulic actuators and load cells. Eliminates effects of nonparallelism, nonperpendicularity, and changes of cable directions upon vector sums of applied forces. Errors due to cable stretching, pulley friction, and weight inaccuracy also eliminated. New instrument rugged and transportable. Set up quickly. Developed to apply known loads to wind-tunnel models with encapsulated strain-gauge balances, also adapted for use in calibrating dynamometers, load sensors on machinery and laboratory instruments.

  8. Procedures for calibration of brachytherapy sources

    International Nuclear Information System (INIS)

    Brachytherapy source strength verification is a responsibility of the user of these source, in fact of the Medical Physicists in charge of this issue in a Radiotherapy Service. The calibration procedures in the users conditions are shown. Specifics methods for source strength determination are recommended, both for High Dose Rate (HDR) sources with Remote Afterloading equipment and for Low Dose Rate sources. The The results of the calibration of HDR Remote After loaders are indicated

  9. Virtual Calibration Chamber CPT on Ticino sand

    OpenAIRE

    Butlanska, Joanna; Arroyo Alvarez de Toledo, Marcos; Gens Solé, Antonio

    2010-01-01

    The following paper surnmarizes results of CPT's performed in virtual calibration chamber (VCC) built with a 3D model based on the distinct element method (DEM). A discrete material tailored to mimic Ticino sand is tested at different densities, stress and stress history. The limit cone tip resistance from the numerical experiments shows quantitative agreement with different empirical curves summarizing previous tests on Ticino sand in physical calibration chambers (ENEL and ISMES).

  10. Calibration of the Super-Kamiokande Detector

    OpenAIRE

    Abe, K.; Hayato, Y.; Iida, T; Iyogi, K.; Kameda, J.; Kishimoto, Y.; Koshio, Y.; Marti, Ll.; Miura, M.; Moriyama, S.; Nakahata, M.(University of Tokyo, Institute for Cosmic Ray Research, Kamioka Observatory, Kamioka, Japan); Nakano, Y.; Nakayama, S.; Obayashi, Y.; Sekiya, H.

    2013-01-01

    Procedures and results on hardware level detector calibration in Super-Kamiokande (SK) are presented in this paper. In particular, we report improvements made in our calibration methods for the experimental phase IV in which new readout electronics have been operating since 2008. The topics are separated into two parts. The first part describes the determination of constants needed to interpret the digitized output of our electronics so that we can obtain physical numbers such as photon count...

  11. Down force calibration stand test report

    International Nuclear Information System (INIS)

    The Down Force Calibration Stand was developed to provide an improved means of calibrating equipment used to apply, display and record Core Sample Truck (CST) down force. Originally, four springs were used in parallel to provide a system of resistance that allowed increasing force over increasing displacement. This spring system, though originally deemed adequate, was eventually found to be unstable laterally. For this reason, it was determined that a new method for resisting down force was needed

  12. Calibration with near-continuous spectral measurements

    DEFF Research Database (Denmark)

    Nielsen, Henrik Aalborg; Rasmussen, Michael; Madsen, Henrik

    2001-01-01

    In chemometrics traditional calibration in case of spectral measurements express a quantity of interest (e.g. a concentration) as a linear combination of the spectral measurements at a number of wavelengths. Often the spectral measurements are performed at a large number of wavelengths and in thi...... by an example in which the octane number of gasoline is related to near infrared spectral measurements. The performance is found to be much better that for the traditional calibration methods....

  13. The Fifth VLBA Calibrator Survey: VCS5

    OpenAIRE

    Kovalev, Y. Y.; Petrov, L.; Fomalont, E. B.; Gordon, D

    2006-01-01

    This paper presents the fifth part of the Very Long Baseline Array (VLBA) Calibrator Survey (VCS), containing 569 sources not observed previously with very long baseline interferometry in geodetic or absolute astrometry programs. This campaign has two goals: (i) to observe additional sources which, together with previous survey results, form a complete sample, (ii) to find new strong sources suitable as phase calibrators. This VCS extension was based on three 24-hour VLBA observing sessions i...

  14. The Sixth VLBA Calibrator Survey: VCS6

    OpenAIRE

    Petrov, L.; Kovalev, Y. Y.; Fomalont, E. B.; Gordon, D

    2008-01-01

    This paper presents the sixth part to the Very Long Baseline Array (VLBA) Calibrator Survey. It contains the positions and maps of 264 sources of which 169 were not previously observed with very long baseline interferometry (VLBI). This survey, based on two 24 hour VLBA observing sessions, was focused on 1) improving positions of 95 sources from previous VLBA Calibrator surveys that were observed either with very large a priori position errors or were observed not long enough to get reliable ...

  15. PFN tool test and calibration system

    International Nuclear Information System (INIS)

    A system has been developed for the functional testing and neutron output calibration of the PFN (Prompt Fission Neutron) Uranium Logging Tool. The system was designed primarily for field work and consists of a special vehicle as well as test apparatus. Only the pertinent instrumentation is described. This document will serve as an Instruction and Test Equipment service manual for those involved with calibration of the neutron output of the PFN tool

  16. Automatic Calibration Of Manual Machine Tools

    Science.gov (United States)

    Gurney, Rex D.

    1990-01-01

    Modified scheme uses data from multiple positions and eliminates tedious positioning. Modification of computer program adapts calibration system for convenient use with manually-controlled machine tools. Developed for use on computer-controlled tools. Option added to calibration program allows data on random tool-axis positions to be entered manually into computer for reduction. Instead of setting axis to predetermined positions, operator merely sets it at variety of arbitrary positions.

  17. MINERνA neutrino detector calibration

    International Nuclear Information System (INIS)

    MINERνA is a neutrino scattering experiment that uses Fermilab’s NuMI beamline. Its goal is to measure cross-sections for neutrino scattering from different nuclei. Precise knowledge of these cross-sections is vital for current and future neutrino oscillation experiments. In order to measure these values to a high degree of accuracy, it is essential that the detector be carefully calibrated. Here, we describe in-situ calibration and cross-checks

  18. Lidar to lidar calibration phase 1

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents a feasibility study of a lidar to lidar (L2L) calibration procedure. Phase one of the project was conducted at Høvsøre, Denmark. Two windcubes were placed next to the 116m met mast and different methods were applied to obtain the sensing height error of the lidars. The purpose...... is to find the most consistent method and use it in a potential lidar to lidar calibration procedure....

  19. Increased Automation in Stereo Camera Calibration Techniques

    OpenAIRE

    Brandi House; Kevin Nickels

    2006-01-01

    Robotic vision has become a very popular field in recent years due to the numerous promising applications it may enhance. However, errors within the cameras and in their perception of their environment can cause applications in robotics to fail. To help correct these internal and external imperfections, stereo camera calibrations are performed. There are currently many accurate methods of camera calibration available; however, most or all of them are time consuming and labor intensive. This r...

  20. Automated Calibration of Multi-Projector Arrays

    OpenAIRE

    Nordbryhn, Ola

    2009-01-01

    Setting up large multi-projector arrays today usually come at a cost; manual calibration of each projector requires time. The orientation of the image from each projector must be correctly aligned in six axes to make the final projected output fit the screen. Not all aspects of calibrating projectors are possible to correct, consumer hardware usually only covers two or three of the axes, the remainder are often corrected using clever projector placement. Also, the degree of which it is possib...

  1. Calibration facility for environment dosimetry instruments

    Energy Technology Data Exchange (ETDEWEB)

    Bercea, Sorin; Celarel, Aurelia; Cenusa, Constantin [Horia Hulubei National Institute for Physics and Nuclear Engineering, 30 Reactorului St, Magurele, Jud Ilfov, P.O.B. MG-6, RO-077125 (Romania)

    2013-12-16

    In the last ten years, the nuclear activities, as well as the major nuclear events (see Fukushima accident) had an increasing impact on the environment, merely by contamination with radioactive materials. The most conferment way to quickly identify the presence of some radioactive elements in the environment, is to measure the dose-equivalent rate H. In this situation, information concerning the values of H due only to the natural radiation background must exist. Usually, the values of H due to the natural radiation background, are very low (∼10{sup −9} - 10{sup −8} Sv/h). A correct measurement of H in this range involve a performing calibration of the measuring instruments in the measuring range corresponding to the natural radiation background lead to important problems due to the presence of the natural background itself the best way to overlap this difficulty is to set up the calibration stand in an area with very low natural radiation background. In Romania, we identified an area with such special conditions at 200 m dept, in a salt mine. This paper deals with the necessary requirements for such a calibration facility, as well as with the calibration stand itself. The paper includes also, a description of the calibration stand (and images) as well as the radiological and metrological parameters. This calibration facilities for environment dosimetry is one of the few laboratories in this field in Europe.

  2. Calibration facility for environment dosimetry instruments

    International Nuclear Information System (INIS)

    In the last ten years, the nuclear activities, as well as the major nuclear events (see Fukushima accident) had an increasing impact on the environment, merely by contamination with radioactive materials. The most conferment way to quickly identify the presence of some radioactive elements in the environment, is to measure the dose-equivalent rate H. In this situation, information concerning the values of H due only to the natural radiation background must exist. Usually, the values of H due to the natural radiation background, are very low (∼10−9 - 10−8 Sv/h). A correct measurement of H in this range involve a performing calibration of the measuring instruments in the measuring range corresponding to the natural radiation background lead to important problems due to the presence of the natural background itself the best way to overlap this difficulty is to set up the calibration stand in an area with very low natural radiation background. In Romania, we identified an area with such special conditions at 200 m dept, in a salt mine. This paper deals with the necessary requirements for such a calibration facility, as well as with the calibration stand itself. The paper includes also, a description of the calibration stand (and images) as well as the radiological and metrological parameters. This calibration facilities for environment dosimetry is one of the few laboratories in this field in Europe

  3. Is Your System Calibrated? MRI Gradient System Calibration for Pre-Clinical, High-Resolution Imaging

    OpenAIRE

    James O'Callaghan; Jack Wells; Simon Richardson; Holly Holmes; Yichao Yu; Simon Walker-Samuel; Bernard Siow; Lythgoe, Mark F

    2014-01-01

    High-field, pre-clinical MRI systems are widely used to characterise tissue structure and volume in small animals, using high resolution imaging. Both applications rely heavily on the consistent, accurate calibration of imaging gradients, yet such calibrations are typically only performed during maintenance sessions by equipment manufacturers, and potentially with acceptance limits that are inadequate for phenotyping. To overcome this difficulty, we present a protocol for gradient calibration...

  4. NuSTAR ground calibration: The Rainwater Memorial Calibration Facility (RaMCaF)

    OpenAIRE

    Brejnholt, Nicolai; Christensen, Finn Erland; Jakobsen, Anders Clemen; Hailey, Charles J.; Koglin, Jason E.; Blaedel, Kenneth L.; Stern, Marcela; Thornhill, Doug; Sleator, Clio; Zhang, Shuo; Craig, William W.; Madsen, Kristin K.; Decker, Todd; Pivovaroff, Michael J.; Vogel, Julia K.

    2011-01-01

    The Nuclear Spectroscopic Telescope Array (NuSTAR) is a NASA Small Explorer mission that will carry the first focusing hard X-ray (5-80 keV ) telescope to orbit. The ground calibration of the three flight optics was carried out at the Rainwater Memorial Calibration Facility (RaMCaF) built for this purpose. In this article we present the facility and its use for the ground calibration of the three optics.

  5. NuSTAR ground calibration: The Rainwater Memorial Calibration Facility (RaMCaF)

    DEFF Research Database (Denmark)

    Brejnholt, Nicolai; Christensen, Finn Erland; Jakobsen, Anders Clemen;

    2011-01-01

    The Nuclear Spectroscopic Telescope Array (NuSTAR) is a NASA Small Explorer mission that will carry the first focusing hard X-ray (5-80 keV ) telescope to orbit. The ground calibration of the three flight optics was carried out at the Rainwater Memorial Calibration Facility (RaMCaF) built for this...... purpose. In this article we present the facility and its use for the ground calibration of the three optics....

  6. Slit-lamp calibration, crucial but neglected

    Directory of Open Access Journals (Sweden)

    Lutfah Rif’ati

    2013-05-01

    Full Text Available AbstrakLatar belakang: Kalibrasi berkala alat diagnostik sangat esensial untuk diagnosis yang akurat. Riset fasilitas kesehatan (Rifaskes 2011 mengumpulkan data termasuk kalibrasi lampu celah (slit-lamp pada sampel rumah sakit (RS di Indonesia. Tujuan analisis ialah untuk mengidentifikasi faktor dominan yang berpengaruh terhadap pelaksanaan kalibrasi berkala lampu celah di RS.Metode: Analisis memakai sebagian data Rifaskes 2011 di antara 442 RS yang menyediakan layanan kesehatan mata. Risiko relatif dipergunakan untuk menilai kemungkinan tidak dilakukannya kalibrasi lampu celah di RS.Hasil: Di antara 248 RS sampel yang memenuhi kriteria inklusi, hanya 25,8% RS yang melakukan kalibrasi lampu celah tepat waktu. Dibandingkan dengan rumah sakit yang dimiliki oleh Badan Usaha Milik Negara (BUMN, rumah sakit yang dimiliki lembaga lain memiliki risiko yang lebih tinggi tidak mengkalibrasi lampu celah. Menurut tipe RS, RS non-pendidikan dibandingkan dengan RS -pendidikan berisiko 40% lebih tinggi tidak mengkalibrasi lampu [risiko relatif suaian (RRa = 1,40; 95% interval kepercayaan (CI = 1,02-1,91].Kesimpulan: Kalibrasi tepat waktu lampu-celah masih menjadi masalah di sebagian besar RS. Dibandingkan dengan rumah sakit yang dimiliki oleh BUMN, rumah sakit yang dimiliki oleh instansi lain berisiko yang lebih tinggi tidak mengkalibrasi lampu celah. (Health Science Indones 2012;2:xx-xxKata kunci:kalibrasi, lampu celah, rumah sakitAbstractBackground: Periodical diagnostic tool calibration is essential for accurate diagnosis. Health Facilities Research (Rifaskes in 2011 collected data on the slit-lamp calibration of all registered general hospitals in Indonesia.Methods: Analysis using a part Rifaskes 2011 data among 442 hospitals that provide eye health services. Relative risk was used to assess the risk of performing calibration slit lamp.Results: Out of 442 hospitals, 248 hospitals met the inclusion study criteria, and only 25.8% calibrating the slit

  7. Facies del subfondo del canal Beagle, Tierra del Fuego

    Directory of Open Access Journals (Sweden)

    Gustavo Bujalesky

    2004-03-01

    Full Text Available El canal Beagle conecta los océanos Pacífico y Atlántico en el extremo meridional de Sudamérica y se ubica en el ambiente subantártico. Conforma una cuenca de unos 300 m de profundidad máxima y está separada del océano Atlántico por un umbral de 30 m de profundidad. El canal es un valle tectónico que fue completamente cubierto por el hielo glacial durante la última glaciación. Posteriormente, el canal fue ocupado por un lago glacial desde los 12.000 a los 8.000 años A.P., cuando fue invadido por el mar que alcanzó un nivel máximo entre los 6.000 y 5.000 años A.P. Con el objetivo de analizar las facies sedimentarias superficiales y del subfondo del canal se realizó un relevamiento geofísico con sonar de barrido lateral y un perfilador de 3,5 kHz. Sobre un basamento constituido por rocas metamórficas del Mesozoico, se identificaron depósitos de till y secuencias granodecrecientes que representan distintos estadios del retroceso glaciar, evidenciando hacia la sección superior facies lacustres y por encima depósitos de la transgresión marina del Holoceno. Además, se han identificado paleocauces y secuencias fluviales cubiertas por sedimentos marinos transgresivos.

  8. New method for the time calibration of an interferometric radio antenna array

    CERN Document Server

    Schröder, F G; Bähren, L; Blümer, J; Bozdog, H; Falcke, H; Haungs, A; Horneffer, A; Huege, T; Isar, P G; Krömer, O; Nehls, S; 10.1016/j.nima.2010.01.072

    2010-01-01

    Digital radio antenna arrays, like LOPES (LOFAR PrototypE Station), detect high-energy cosmic rays via the radio emission from atmospheric extensive air showers. LOPES is an array of dipole antennas placed within and triggered by the KASCADE-Grande experiment on site of the Karlsruhe Institute of Technology, Germany. The antennas are digitally combined to build a radio interferometer by forming a beam into the air shower arrival direction which allows measurements even at low signal-to-noise ratios in individual antennas. This technique requires a precise time calibration. A combination of several calibration steps is used to achieve the necessary timing accuracy of about 1 ns. The group delays of the setup are measured, the frequency dependence of these delays (dispersion) is corrected in the subsequent data analysis, and variations of the delays with time are monitored. We use a transmitting reference antenna, a beacon, which continuously emits sine waves at known frequencies. Variations of the relative del...

  9. Estudio del proceso de crecimiento del cultivo del tomate

    OpenAIRE

    Barraza, Fernando V.; Fischer, Gerhard; Cardona, Carlos E.

    2010-01-01

    Para conocer la densidad poblacional óptimay obtener el mayor rendimiento en el cultivo deltomate en el Valle del Sinú medio, se hizo una investigaciónmediante diseño completamente al azar con cuatrotratamientos (densidades poblacionales de 20.000,25.000, 33.333 y 50.000 plantas∙ha -1) y tres repeticiones.Se tomaron las medidas directas: altura de planta, númerode hojas, área foliar, número de frutos y rendimiento.También se calcularon los índices de crecimiento: tasa decrecimiento del cultiv...

  10. LINEAR AND NON-LINEAR CAMERA CALIBRATION TECHNIQUES

    OpenAIRE

    Manoj Gupta

    2011-01-01

    This Paper deals with calibrate a camera to find out the intrinsic and extrinsic camera parameters which are necessary to recover the depth estimation of an object in stereovision system. Keywords: Camera Calibration, Tsai’s algorithm, Stereovision, Linear Calibration, Non-Linear Calibration, Depth estimation

  11. Calibration Of Partial-Pressure-Of-Oxygen Sensors

    Science.gov (United States)

    Yount, David W.; Heronimus, Kevin

    1995-01-01

    Report and analysis of, and discussion of improvements in, procedure for calibrating partial-pressure-of-oxygen sensors to satisfy Spacelab calibration requirements released. Sensors exhibit fast drift, which results in short calibration period not suitable for Spacelab. By assessing complete process of determining total drift range available, calibration procedure modified to eliminate errors and still satisfy requirements without compromising integrity of system.

  12. Ørsted Pre-Flight Magnetometer Calibration Mission

    DEFF Research Database (Denmark)

    Risbo, T.; Brauer, Peter; Merayo, José M.G.; Nielsen, Otto V; Petersen, Jan Raagaard; Primdahl, Fritz

    2003-01-01

    overall calibration results are given. The temperature calibrations are explained and reported on. The overall calibration model standard deviation is about 100 pT rms. Comparisons with the later in-flight calibrations show that, except for the unknown satellite offsets, an agreement within 4 nT was...

  13. 40 CFR 1065.330 - Exhaust-flow calibration.

    Science.gov (United States)

    2010-07-01

    ... recommend that you use a calibration subsonic venturi or ultrasonic flow meter and simulate exhaust... ultrasonic flow meter for raw exhaust flow measurement, we recommend that you calibrate it as described in... Exhaust-flow calibration. (a) Calibrate exhaust-flow meters upon initial installation. Follow...

  14. Programmable and automatic calibrator for radio sources at 45 MHz

    Science.gov (United States)

    Aparici, J.; May, J.; Salas, F.; Ventura, J.

    1981-12-01

    The design, construction and operation of a standard calibrator is presented. The calibrator consists of saturated diodes controlled by an indirect feed-back system and a digital-to-analog converter. The advantages over similar designs are described, as for instance, high-resolution in the calibration scale, good stability, very fast calibrations, use of balanced electronic switches, etc.

  15. 21 CFR 874.1080 - Audiometer calibration set.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Audiometer calibration set. 874.1080 Section 874...) MEDICAL DEVICES EAR, NOSE, AND THROAT DEVICES Diagnostic Devices § 874.1080 Audiometer calibration set. (a) Identification. An audiometer calibration set is an electronic reference device that is intended to calibrate...

  16. Adaptable Multivariate Calibration Models for Spectral Applications

    Energy Technology Data Exchange (ETDEWEB)

    THOMAS,EDWARD V.

    1999-12-20

    Multivariate calibration techniques have been used in a wide variety of spectroscopic situations. In many of these situations spectral variation can be partitioned into meaningful classes. For example, suppose that multiple spectra are obtained from each of a number of different objects wherein the level of the analyte of interest varies within each object over time. In such situations the total spectral variation observed across all measurements has two distinct general sources of variation: intra-object and inter-object. One might want to develop a global multivariate calibration model that predicts the analyte of interest accurately both within and across objects, including new objects not involved in developing the calibration model. However, this goal might be hard to realize if the inter-object spectral variation is complex and difficult to model. If the intra-object spectral variation is consistent across objects, an effective alternative approach might be to develop a generic intra-object model that can be adapted to each object separately. This paper contains recommendations for experimental protocols and data analysis in such situations. The approach is illustrated with an example involving the noninvasive measurement of glucose using near-infrared reflectance spectroscopy. Extensions to calibration maintenance and calibration transfer are discussed.

  17. Calibration Matters: Advances in Strapdown Airborne Gravimetry

    Science.gov (United States)

    Becker, D.

    2015-12-01

    Using a commercial navigation-grade strapdown inertial measurement unit (IMU) for airborne gravimetry can be advantageous in terms of cost, handling, and space consumption compared to the classical stable-platform spring gravimeters. Up to now, however, large sensor errors made it impossible to reach the mGal-level using such type IMUs as they are not designed or optimized for this kind of application. Apart from a proper error-modeling in the filtering process, specific calibration methods that are tailored to the application of aerogravity may help to bridge this gap and to improve their performance. Based on simulations, a quantitative analysis is presented on how much IMU sensor errors, as biases, scale factors, cross couplings, and thermal drifts distort the determination of gravity and the deflection of the vertical (DOV). Several lab and in-field calibration methods are briefly discussed, and calibration results are shown for an iMAR RQH unit. In particular, a thermal lab calibration of its QA2000 accelerometers greatly improved the long-term drift behavior. Latest results from four recent airborne gravimetry campaigns confirm the effectiveness of the calibrations applied, with cross-over accuracies reaching 1.0 mGal (0.6 mGal after cross-over adjustment) and DOV accuracies reaching 1.1 arc seconds after cross-over adjustment.

  18. Evolution of Altimetry Calibration and Future Challenges

    Science.gov (United States)

    Fu, Lee-Lueng; Haines, Bruce J.

    2012-01-01

    Over the past 20 years, altimetry calibration has evolved from an engineering-oriented exercise to a multidisciplinary endeavor driving the state of the art. This evolution has been spurred by the developing promise of altimetry to capture the large-scale, but small-amplitude, changes of the ocean surface containing the expression of climate change. The scope of altimeter calibration/validation programs has expanded commensurately. Early efforts focused on determining a constant range bias and verifying basic compliance of the data products with mission requirements. Contemporary investigations capture, with increasing accuracies, the spatial and temporal characteristics of errors in all elements of the measurement system. Dedicated calibration sites still provide the fundamental service of estimating absolute bias, but also enable long-term monitoring of the sea-surface height and constituent measurements. The use of a network of island and coastal tide gauges has provided the best perspective on the measurement stability, and revealed temporal variations of altimeter measurement system drift. The cross-calibration between successive missions provided fundamentally new information on the performance of altimetry systems. Spatially and temporally correlated errors pose challenges for future missions, underscoring the importance of cross-calibration of new measurements against the established record.

  19. Computerized 50 liter volume calibration system

    International Nuclear Information System (INIS)

    A system has been designed for the Savannah River Site that will be used to calibrate product shipping containers. For accountability purposes, it is necessary that these containers be calibrated to a very high precision. The Computerized 50 Liter Volume Calibration System (CVCS), which is based on the Ideal Gas Law (IGL), will use reference volumes with precision of no less ±0.03%, and helium to calibrate the containers to have a total error of no greater than ±0.10%. A statistical interpretation of the system has given a theoretical total calculated error of ±0.08%. Tests with the system will be performed once fabrication is complete to experimentally verify the calculated error. Since the total error was calculated using the worst case scenario, the actual error should be significantly less than the calculated value. The computer controlled, totally automated system is traceable to the National Institute of Standards and Technology. The design, calibration procedure, and statistical interpretation of the system will be discussed. 1 ref

  20. PERSONALISED BODY COUNTER CALIBRATION USING ANTHROPOMETRIC PARAMETERS.

    Science.gov (United States)

    Pölz, S; Breustedt, B

    2016-09-01

    Current calibration methods for body counting offer personalisation for lung counting predominantly with respect to ratios of body mass and height. Chest wall thickness is used as an intermediate parameter. This work revises and extends these methods using a series of computational phantoms derived from medical imaging data in combination with radiation transport simulation and statistical analysis. As an example, the method is applied to the calibration of the In Vivo Measurement Laboratory (IVM) at Karlsruhe Institute of Technology (KIT) comprising four high-purity germanium detectors in two partial body measurement set-ups. The Monte Carlo N-Particle (MCNP) transport code and the Extended Cardiac-Torso (XCAT) phantom series have been used. Analysis of the computed sample data consisting of 18 anthropometric parameters and calibration factors generated from 26 photon sources for each of the 30 phantoms reveals the significance of those parameters required for producing an accurate estimate of the calibration function. Body circumferences related to the source location perform best in the example, while parameters related to body mass show comparable but lower performances, and those related to body height and other lengths exhibit low performances. In conclusion, it is possible to give more accurate estimates of calibration factors using this proposed approach including estimates of uncertainties related to interindividual anatomical variation of the target population. PMID:26396263

  1. Numerical calibration of a Lorentz force flowmeter

    International Nuclear Information System (INIS)

    Numerical simulation of complex-shaped devices for contactless electromagnetic flow measurement in metallurgy is a challenge for computational magnetohydrodynamics. We report a series of numerical simulations which demonstrate for the first time that it is possible to predict the calibration constant of a generic Lorentz force flowmeter (LFF) with an uncertainty close to the requirements of real-life industrial applications. Our simulations involve both magnetostatic computations of a complex-shaped magnet system and magnetohydrodynamic computations of the flow of a liquid metal in a nozzle under the influence of a predominantly transverse magnetic field. In order to assess the role of turbulence, the simulations have been performed both for laminar and for turbulent flows using Reynolds-averaged Navier–Stokes equations in the latter case. In addition to the numerical simulations we have measured the calibration constant of the considered LFF using room-temperature liquid metal instead of liquid aluminum. A comparison between the numerically predicted and the measured values of the calibration constant shows that they differ by only 3.4%. This result suggests that numerical calibration of a LFF may become an economic alternative to expensive full-scale experimental calibration. (paper)

  2. Calibration of the ZEUS forward calorimeter

    International Nuclear Information System (INIS)

    The physics at the ep-collider HERA requires high resolution calorimetry calibrated with an accuracy of better than 2%. The ZEUS detector meets these conditions by means of a compensating uranium scintillator sandwich calorimeter with an energy resolution of σ/E = 35%/√E + σcal, where σcal is the calibration error. One of the tools to minimize σcal is the calibration with the signals of the radioactivity of the Uranium plates (UNO). Taking UNO data every 8 hours keeps the calibration stable within ≅ 1%. The muon calibration is done employing an algorithm, that determines the most probable energy loss with a precision of ≅ 1%. The channel-to-channel fluctuations of the ratio μ/UNO for a forward calorimeter (FCAL) prototype show a spread of 5.2% for the electromagnetic calorimeter and ≅ 2.5% for the hadronic sections. Improvements in the construction of the FCAL modules decreased these fluctuations to 2.0% and ≅ 1.8% respectively. The influence of the cracks between the calorimeter modules amounts to ≅ 1.7% on average for the ZEUS geometry, if a 2 mm thick Pb-sheet is introduced between the modules. We conclude that we are able to keep σcal below 2%. (orig.)

  3. WFC3: Improved WFC3 Calibration Products

    Science.gov (United States)

    Gunning, Heather C.; Sosey, M. L.; Anderson, J.; Lee, J. C.; Pirzkal, N.; MacKenty, J. W.; Kozhurina-Platais, V.; Deustua, S. E.; Hammer, D.; Dahlen, T.; Sabbi, E.; Mack, J.; Baggett, S. M.; WFC3 Team

    2014-01-01

    The Wide Field Camera 3 (WFC3) is a fourth-generation UV/visible and IR imaging instrument on the Hubble Space Telescope (HST). Installed in May 2009, during HST servicing mission 4, both channels have been performing very well on-orbit. To provide optimum calibrated data, the WFC3 team routinely updates and refines the calibration software and associated files, designated as calibration products. We present some of the recently improved calibration products that will be of interest to current and future users of WFC3, including information on the chip-dependent zeropoints and flat fields, post-flash calibrations, and detector-to-image distortion corrections. The latter results in four new extensions (two per chip and dimension), in all UVIS FLTs retrieved from MAST after September 10, 2013. The D2IMFILE contains astrometric corrections for shifts of the raw X and Y positions induced by the lithographic-mask pattern. We discuss the migration of CALWF3 from the STSDAS package to HSTCAL, a package independent of IRAF; as a consequence, the IRAF/STSDAS version of CALWF3 is no longer being updated. Finally, we summarize recent improvements to aXe, a PyRAF/IRAF software package that enables automated extraction of spectra from WFC3 slitless spectral (grism) images. Updated versions of aXe are made available as part of the STSDAS testing environment (SSBX).

  4. Marketing Mix del Software.

    OpenAIRE

    Yudith del Carmen Rodríguez Pérez

    2006-01-01

    La ingeniería del software y los modelos de calidad del software han consolidado sus esfuerzos en el proceso de producción del mismo, sin embargo son pocos sus aportes en el proceso de comercialización. Es esencial en la ciencia de la computación desarrollar un modelo de comercialización para las organizaciones productoras de software con el fin de elevar la productividad de las mismas. Sin embargo, es preciso primero conocer las características del producto software que los diferencian de ot...

  5. Los sentimientos del Cid

    OpenAIRE

    Aurelio González

    2007-01-01

    Un recorrido por los sentimientos que se han plasmado en la caracterización del personaje del Cid conlleva el análisis de las formas expresivas, que se utilizan para que estos sentimientos sean verosímiles, sin entrar en contradicción con la condición heroica y épica del personaje. La consideración de las expresiones de tristeza, alegría, devoción o gratitud de Rodrigo permite identificar estrategias discursivas del poema que intentan acercar al receptor a una valoración positiva de la baja n...

  6. Least-Squares Camera Calibration Including Lens Distortion and Automatic Editing of Calibration Points

    Science.gov (United States)

    Gennery, D. B.

    1998-01-01

    A method is described for calibrating cameras including radial lens distortion, by using known points such as those measured from a calibration fixture. The distortion terms are relative to the optical axis, which is included in the model so that it does not have to be orthogonal to the image sensor plane.

  7. Estudio del léxico del Ecuador

    OpenAIRE

    Estrella Santos, Ana Teresa

    2008-01-01

    El presente trabajo de investigación pretende proporcionar una descripción del léxico del español del Ecuador, a partir de materiales de la lengua hablada. Hemos llevado a cabo un análisis contrastivo del léxico de las provincias de Pichincha y Guayas, una recopilación del léxico general del Ecuador y un análisis de los quichuismos presentes en nuestro corpus.

  8. Riesgos laborales en trabajadores del sector informal del Cauca, Colombia

    OpenAIRE

    Augusto Muñoz-Caicedo; Pilar Mirely Chois-Lenis

    2014-01-01

    Antecedentes. Un estudio a nivel nacional permitió establecer los riesgos a los cuales estaban expuestos los trabajadores del sector informal del comercio. Sin embargo, no fue incluido el Departamento del Cauca. Objetivo. Describirlos riesgos laborales de las personas ocupadas en el sector informal del comercio, en cinco municipios del departamento del Cauca, durante el 2011. Materiales y métodos. Estudio descriptivo, transversal, desarrollado con 223 sujetos a quienes se les aplicó una encue...

  9. An Optimal Calibration Method for a MEMS Inertial Measurement Unit

    Directory of Open Access Journals (Sweden)

    Bin Fang

    2014-02-01

    Full Text Available An optimal calibration method for a micro-electro-mechanical inertial measurement unit (MIMU is presented in this paper. The accuracy of the MIMU is highly dependent on calibration to remove the deterministic errors of systematic errors, which also contain random errors. The overlapping Allan variance is applied to characterize the types of random error terms in the measurements. The calibration model includes package misalignment error, sensor-to-sensor misalignment error and bias, and a scale factor is built. The new concept of a calibration method, which includes a calibration scheme and a calibration algorithm, is proposed. The calibration scheme is designed by D-optimal and the calibration algorithm is deduced by a Kalman filter. In addition, the thermal calibration is investigated, as the bias and scale factor varied with temperature. The simulations and real tests verify the effectiveness of the proposed calibration method and show that it is better than the traditional method.

  10. A method of calibrating wind velocity sensors with a modified gas flow calibrator

    Science.gov (United States)

    Stump, H. P.

    1978-01-01

    A procedure was described for calibrating air velocity sensors in the exhaust flow of a gas flow calibrator. The average velocity in the test section located at the calibrator exhaust was verified from the mass flow rate accurately measured by the calibrator's precision sonic nozzles. Air at elevated pressures flowed through a series of screens, diameter changes, and flow straighteners, resulting in a smooth flow through the open test section. The modified system generated air velocities of 2 to 90 meters per second with an uncertainty of about two percent for speeds below 15 meters per second and four percent for the higher speeds. Wind tunnel data correlated well with that taken in the flow calibrator.

  11. e-Calibrations: using the Internet to deliver calibration services in real time at lower cost

    International Nuclear Information System (INIS)

    The National Institute of Standards and Technology (NIST) is expanding into a new frontier in the delivery of measurement services. The Internet will be employed to provide industry with electronic traceability to national standards. This is a radical departure from the traditional modes of traceability and presents many new challenges. The traditional mail-based calibration service relies on sending artifacts to the user, who then mails them back to NIST for evaluation. The new service will deliver calibration results to the industry customer on-demand, in real-time, at a lower cost. The calibration results can be incorporated rapidly into the production process to ensure the highest quality manufacturing. The service would provide the US radiation processing industry with a direct link to the NIST calibration facilities and its expertise, and provide an interactive feedback process between industrial processing and the national measurement standard. Moreover, an Internet calibration system should contribute to the removal of measurement-related trade barriers

  12. Syringe calibration factors and volume correction factors for the NPL secondary standard radionuclide calibrator

    CERN Document Server

    Tyler, D K

    2002-01-01

    The activity assay of a radiopharmaceutical administration to a patient is normally achieved via the use of a radionuclide calibrator. Because of the different geometries and elemental compositions between plastic syringes and glass vials, the calibration factors for syringes may well be significantly different from those for the glass containers. The magnitude of these differences depends on the energies of the emitted photons. For some radionuclides variations have been observed of 70 %, it is therefore important to recalibrate for syringes or use syringe calibration factors. Calibration factors and volume correction factors have been derived for the NPL secondary standard radionuclide calibrator, for a variety of commonly used syringes and needles, for the most commonly used medical radionuclide.

  13. A Review of Sensor Calibration Monitoring for Calibration Interval Extension in Nuclear Power Plants

    Energy Technology Data Exchange (ETDEWEB)

    Coble, Jamie B.; Meyer, Ryan M.; Ramuhalli, Pradeep; Bond, Leonard J.; Hashemian, Hash; Shumaker, Brent; Cummins, Dara

    2012-08-31

    Currently in the United States, periodic sensor recalibration is required for all safety-related sensors, typically occurring at every refueling outage, and it has emerged as a critical path item for shortening outage duration in some plants. Online monitoring can be employed to identify those sensors that require calibration, allowing for calibration of only those sensors that need it. International application of calibration monitoring, such as at the Sizewell B plant in United Kingdom, has shown that sensors may operate for eight years, or longer, within calibration tolerances. This issue is expected to also be important as the United States looks to the next generation of reactor designs (such as small modular reactors and advanced concepts), given the anticipated longer refueling cycles, proposed advanced sensors, and digital instrumentation and control systems. The U.S. Nuclear Regulatory Commission (NRC) accepted the general concept of online monitoring for sensor calibration monitoring in 2000, but no U.S. plants have been granted the necessary license amendment to apply it. This report presents a state-of-the-art assessment of online calibration monitoring in the nuclear power industry, including sensors, calibration practice, and online monitoring algorithms. This assessment identifies key research needs and gaps that prohibit integration of the NRC-approved online calibration monitoring system in the U.S. nuclear industry. Several needs are identified, including the quantification of uncertainty in online calibration assessment; accurate determination of calibration acceptance criteria and quantification of the effect of acceptance criteria variability on system performance; and assessment of the feasibility of using virtual sensor estimates to replace identified faulty sensors in order to extend operation to the next convenient maintenance opportunity. Understanding the degradation of sensors and the impact of this degradation on signals is key to

  14. ATLAS Tile calorimeter calibration and monitoring systems

    CERN Document Server

    Chomont, Arthur Rene; The ATLAS collaboration

    2016-01-01

    The ATLAS Tile Calorimeter (TileCal) is the central section of the hadronic calorimeter of the ATLAS experiment and provides important information for reconstruction of hadrons, jets, hadronic decays of tau leptons and missing transverse energy. This sampling calorimeter uses steel plates as absorber and scintillating tiles as active medium. The light produced by the passage of charged particles is transmitted by wavelength shifting fibres to photomultiplier tubes (PMTs), located on the outside of the calorimeter. The readout is segmented into about 5000 cells (longitudinally and transversally), each of them being read out by two PMTs in parallel. To calibrate and monitor the stability and performance of each part of the readout chain during the data taking, a set of calibration systems is used. The TileCal calibration system comprises Cesium radioactive sources, laser and charge injection elements and it allows to monitor and equalize the calorimeter response at each stage of the signal production, from scin...

  15. Calibration of fisheye lenses for hemispherical photography

    International Nuclear Information System (INIS)

    Hemispherical photography represents one of the most appropriate methods of estimating averages of solar radiation over extended periods of time. This method is based upon the use of extremely wide-angle fisheye lenses, which produce large projection distortion. To correctly interpret hemispherical photography we have to know the projection characteristics of the fisheye lens in combination with a camera body. This can be achieved through lens calibration. The first part of the article explains in detail the calibration method for fisheye lenses which are used to assess the solar radiation in forest ecology research. In the second part the results of calibration for fisheye lens Sigma 8 mm, f/4 (MF, N) are presented. The lens was used on a Nikon F50 camera body

  16. Oregon State TRIGA reactor power calibration study

    International Nuclear Information System (INIS)

    As a result of a recent review of the Oregon State TRIGA Reactor (OSTR) power calibration procedure, an investigation was performed on the origin and correctness of the OSTR tank factor and the calibration method. It was determined that there was no clear basis for the tank factor which was being used (0.0525 deg. C/kwh) and therefore a new value was calculated (0.0493 deg. C/kwh). The calculational method and likely errors are presented in the paper. In addition, a series of experimental tests were conducted to decide if the power calibration was best performed with or without a mixer, at 100 KW or at 1 MW. The results of these tests along with the final recommendation are presented. (author)

  17. Ion Beam Energy Calibration Method for Accelerator

    International Nuclear Information System (INIS)

    Ion beam energy calibration methods, i e : nuclear reaction method, magnetic field method and calorimeter method were elaborated and studied from its advantage and disadvantage in this paper. Ion beam energy calibration method for accelerator using the method of magnetic field on 3 MV Tandem Accelerator have been carried out at Tiara, JAERI, Japan. The result showed that the energy of ion beam current is 43.56 keV. The result of study conclude that nuclear reaction method generally used to calibrate ion beam energy at the accelerator of energy larger than 2 MeV, calorimetric method for the accelerator electron including linac, magnetic field method for all particle type of accelerator. (author)

  18. The Calibration of the FACT Camera

    International Nuclear Information System (INIS)

    Full text: The First G-APD Cherenkov Telescope (FACT) collaboration builds a camera for an Imaging Atmospheric Cherenkov Telescope which is based on G-APDs and a readout using the Domino Ring Sampling (DRS4) chip. The amplitude calibration of the readout chain must account for a wide variety of effects specific to this design of the camera, eg. the strong temperature dependence of the G-APDs, the quality of the gluing between the optical components as well as the characteristics of the DRS4 chip. The basis for this calibration are an online feedback system to stabilize the gain of the G-APDs, laboratory measurements and special runs during data taking. In this talk, the calibration system for FACT is presented including the current experience with the camera in laboratory measurements. (author)

  19. Muon Calibration at SoLid

    CERN Document Server

    Saunders, Daniel

    2016-01-01

    The SoLid experiment aims to make a measurement of very short distance neutrino oscillations using reactor antineutrinos. Key to its sensitivity are the experiments high spatial and energy resolution, combined with a very suitable reactor source and efficient background rejection. The fine segmentation of the detector (cubes of side 5cm), and ability to resolve signals in space and time, gives SoLid the capability to track cosmic muons. In principle a source of background, these turn into a valuable calibration source if they can be cleanly identified. This work presents the first energy calibration results, using cosmic muons, of the 288kg SoLid prototype SM1. This includes the methodology of tracking at SoLid, cosmic ray angular analyses at the reactor site, estimates of the time resolution, and calibrations at the cube level.

  20. Photometric Calibrations for 21st Century Science

    CERN Document Server

    Kent, Stephen; Deustua, Susana E; Smith, J Allyn; Adelman, Saul; Allam, Sahar; Baptista, Brian; Bohlin, Ralph C; Clem, James L; Conley, Alex; Edelstein, Jerry; Elias, Jay; Glass, Ian; Henden, Arne; Howell, Steve; Kimble, Randy A; Kruk, Jeffrey W; Lampton, Michael; Magnier, Eugene A; McCandliss, Stephan R; Moos, Warren; Mostek, Nick; Mufson, Stuart; Oswalt, Terry D; Perlmutter, Saul; Prieto, Carlos Allende; Rauscher, Bernard J; Riess, Adam; Saha, Abhijit; Sullivan, Mark; Suntzeff, Nicholas; Tokunaga, Alan; Tucker, Douglas; Wing, Robert; Woodgate, Bruce; Wright, Edward L

    2009-01-01

    The answers to fundamental science questions in astrophysics, ranging from the history of the expansion of the universe to the sizes of nearby stars, hinge on our ability to make precise measurements of diverse astronomical objects. As our knowledge of the underlying physics of objects improves along with advances in detectors and instrumentation, the limits on our capability to extract science from measurements is set, not by our lack of understanding of the nature of these objects, but rather by the most mundane of all issues: the precision with which we can calibrate observations in physical units. We stress the need for a program to improve upon and expand the current networks of spectrophotometrically calibrated stars to provide precise calibration with an accuracy of equal to and better than 1% in the ultraviolet, visible and near-infrared portions of the spectrum, with excellent sky coverage and large dynamic range.

  1. Calibration of mixed-polarization interferometric observations

    CERN Document Server

    Marti-Vidal, Ivan; Conway, John; Zensus, Anton J

    2016-01-01

    Heterodyne receivers register the sky signal on either a circular polarization basis (where it is split into left-hand and right-hand circular polarization) or a linear polarization basis (where it is split into horizontal and vertical linear polarization). We study the problem of interferometric observations performed with telescopes that observe on different polarization bases, hence producing visibilities that we call "mixed basis" (i.e., linear in one telescope and circular in the other). We present novel algorithms for the proper calibration and treatment of such interferometric observations and test our algorithms with both simulations and real data. The use of our algorithms will be important for the optimum calibration of forthcoming observations with the Atacama Large mm/submm Array (ALMA) in very-long-baseline interferometry (VLBI) mode. Our algorithms will also allow us to optimally calibrate future VLBI observations at very high data rates (i.e., wide bandwidths), where linear-polarization feeds w...

  2. Multichannel radiometer calibration: a new approach

    Science.gov (United States)

    Diaz, Susana; Booth, Charles R.; Armstrong, Roy; Brunat, Claudio; Cabrera, Sergio; Camilion, Carolina; Casiccia, Claudio; Deferrari, Guillermo; Fuenzalida, Humberto; Lovengreen, Charlotte; Paladini, Alejandro; Pedroni, Jorge; Rosales, Alejandro; Zagarese, Horacio; Vernet, Maria

    2005-09-01

    The error in irradiance measured with Sun-calibrated multichannel radiometers may be large when the solar zenith angle (SZA) increases. This could be particularly detrimental in radiometers installed at mid and high latitudes, where SZAs at noon are larger than 50° during part of the year. When a multiregressive methodology, including the total ozone column and SZA, was applied in the calculation of the calibration constant, an important improvement was observed. By combining two different equations, an improvement was obtained at almost all the SZAs in the calibration. An independent test that compared the irradiance of a multichannel instrument and a spectroradiometer installed in Ushuaia, Argentina, was used to confirm the results.

  3. The Fourth VLBA Calibrator Survey - VCS4

    CERN Document Server

    Petrov, L A; Fomalont, E B; Gordon, D

    2006-01-01

    This paper presents the fourth extension to the Very Long Baseline Array (VLBA) Calibrator Survey, containing 262 new sources not previously observed with very long baseline interferometry (VLBI). This survey, based on three 24 hour VLBA observing sessions, fills remaining areas on the sky above declination -40 degrees where the calibrator density is less than one source within a 4 degree radius disk at any given direction. The share of these area was reduced from 4.6% to 1.9%. Source positions were derived from astrometric analysis of group delays determined at 2.3 and 8.6 GHz frequency bands using the Calc/Solve software package. The VCS4 catalogue of source positions, plots of correlated flux density versus projected baseline length, contour plots and fits files of naturally weighted CLEAN images, as well as calibrated visibility function files are available on the Web at http://gemini.gsfc.nasa.gov/vcs4 .

  4. MAGNETIC GRADIOMETRY: Instrumentation, Calibration and Applications

    DEFF Research Database (Denmark)

    Merayo, Jose Maria Garcia

    The description of the single axis magnetic gradiometer based on two Compensation Detector Coil (CDC) fluxgate ringcore sensors separated 20cm introduces the subject of magnetic gradiometry. Despite its good properties and high precision of less than 1nT, the calibration procedures...... absolute accuracy of 93pT/m.The scalar calibration of a vector magnetometer is explained thoroughly. The novel method is simple and it represents the most robust and unique way to estimate the characterizing 9 parameters of a vector magnetometer. Its power relies on the linearization of the parametrization...... by comparing the square of the intensities of the reference and of the uncalibrated magnetometers in the Earth's magnetic field. Using this method a CSC magnetometer can be absolutely calibrated with 0.2nT of accuracy.The absolute alignment of a vector magnetometer is also described. After the scalar...

  5. Implementing new recommendations for calibrating personal dosemeters

    International Nuclear Information System (INIS)

    This paper analyses the differences between the calibration procedures for personal dosemeters recommended by ICRU 47 and ISO 4037-3. The tissue equivalence of the PMMA and the ISO water slab phantoms are analysed by means of the Penelope Monte Carlo code for monoenergetic and filtered X ray photon beams and compared with the results of two other independent codes. The influence of the calibration method is also verified experimentally, both on a thermoluminescence and an electronic personal dosemeter. Good consistency between both calibration procedures is shown provided that a correction factor for backscatter differences between the PMMA and the ICRU phantom is introduced. The Monte Carlo simulation is used to determine this correction to a greater accuracy. (author)

  6. Control console of the gamma calibration room

    International Nuclear Information System (INIS)

    The Nuclear Centre of Mexico has a Ionizing Radiation Metrology Center (CMRI). This is in charge of the calibration in Mexico and Latin America of equipment dedicated to radiation measurement as industrial, medical as other fields. The importance to ensure that the equipment stay justly calibrated, it is imposed the necessity of automating the different rooms which the CMRI has. in this case it will be exposed the Calibration room for gamma radiation type. The operation of this application was carried out with the LabVIEW development platform and also in C language. The hardware associated is: personal computer with two cards using the 8255 device, 16 channels with optical isolation to manage input/output TTL type, 16 channels with optical isolation to management of charges to 127 V a.c., and 2 channels for 90V d.c. motors. (Author)

  7. Kinematic calibration of orthoglide-type mechanisms

    CERN Document Server

    Pashkevich, Anatoly; Wenger, Philippe

    2006-01-01

    The paper proposes a novel calibration approach for the Orthoglide-type mechanisms based on observations of the manipulator leg parallelism during mo-tions between the prespecified test postures. It employs a low-cost measuring system composed of standard comparator indicators attached to the universal magnetic stands. They are sequentially used for measuring the deviation of the relevant leg location while the manipulator moves the TCP along the Cartesian axes. Using the measured differences, the developed algorithm estimates the joint offsets that are treated as the most essential parameters to be adjusted. The sensitivity of the meas-urement methods and the calibration accuracy are also studied. Experimental re-sults are presented that demonstrate validity of the proposed calibration technique

  8. Mobile Camera Array Calibration for Light Field Acquisition

    OpenAIRE

    Xu, Yichao; Maeno, Kazuki; Nagahara, Hajime; Taniguchi, Rin-ichiro

    2014-01-01

    The light field camera is useful for computer graphics and vision applications. Calibration is an essential step for these applications. After calibration, we can rectify the captured image by using the calibrated camera parameters. However, the large camera array calibration method, which assumes that all cameras are on the same plane, ignores the orientation and intrinsic parameters. The multi-camera calibration technique usually assumes that the working volume and viewpoints are fixed. In ...

  9. The cryogenic balance design and balance calibration methods

    Science.gov (United States)

    Ewald, B.; Polanski, L.; Graewe, E.

    1992-07-01

    The current status of a program aimed at the development of a cryogenic balance for the European Transonic Wind Tunnel is reviewed. In particular, attention is given to the cryogenic balance design philosophy, mechanical balance design, reliability and accuracy, cryogenic balance calibration concept, and the concept of an automatic calibration machine. It is shown that the use of the automatic calibration machine will improve the accuracy of calibration while reducing the man power and time required for balance calibration.

  10. Local Volatility Calibration Using An Adjoint Proxy

    Directory of Open Access Journals (Sweden)

    Gabriel TURINICI

    2008-11-01

    Full Text Available We document the calibration of the local volatility in a framework similar to Coleman, Li and Verma. The quality of a surface is assessed through a functional to be optimized; the specificity of the approach is to separate the optimization (performed with any suitable optimization algorithm from the computation of the functional where we use an adjoint (as in L. Jiang et. al. to obtain an approximation; moreover our main calibration variable is the implied volatility (the procedure can also accommodate the Greeks. The procedure performs well on benchmarks from the literature and on FOREX data.

  11. Geometric calibration of ERS satellite SAR images

    DEFF Research Database (Denmark)

    Mohr, Johan Jacob; Madsen, Søren Nørvang

    2001-01-01

    Geometric calibration of the European Remote Sensing (ERS) Satellite synthetic aperture radar (SAR) slant range images is important in relation to mapping areas without ground reference points and also in relation to automated processing. The relevant SAR system parameters are discussed and...... calibrated by using the European Space Agency (ESA) transponders at Flevoland. The resulting accuracy of the slant range images corresponds to 10 m horizontally on the ground. The results are verified by using runway intersections and corner reflectors surveyed with differential GPS techniques. Based on a...

  12. HENC performance evaluation and plutonium calibration

    International Nuclear Information System (INIS)

    The authors have designed a high-efficiency neutron counter (HENC) to increase the plutonium content in 200-L waste drums. The counter uses totals neutron counting, coincidence counting, and multiplicity counting to determine the plutonium mass. The HENC was developed as part of a Cooperative Research and Development Agreement between the Department of Energy and Canberra Industries. This report presents the results of the detector modifications, the performance tests, the add-a-source calibration, and the plutonium calibration at Los Alamos National Laboratory (TA-35) in 1996

  13. Electrical Calibration for Calvet Type Microcalorimetry

    Institute of Scientific and Technical Information of China (English)

    2001-01-01

    The deyjce construction and working principle of Calvet type heat conductive microcalorimetry are generally ineroduced in this paper. Based on previous work, we design a reactor vessel and its electrical calibration equipment according to our special study systems. The heat constant of our improved microalorimetry is gotten by means of electrical calibration on the basis of Tian's equation. The results gotten by in tegration method are identical with ones gotten by peak method. It shows that our improved device works well enough to develop further studies.``

  14. Time calibration of the ANTARES neutrino telescope

    International Nuclear Information System (INIS)

    The ANTARES deep-sea neutrino telescope comprises a three-dimensional array of photomultipliers to detect the Cherenkov light induced by upgoing relativistic charged particles originating from neutrino interactions in the vicinity of the detector. The large scattering length of light in the deep sea facilitates an angular resolution of a few tenths of a degree for neutrino energies exceeding 10 TeV. In order to achieve this optimal performance, the time calibration procedures should ensure a relative time calibration between the photomultipliers at the level of 1 ns. The methods developed to attain this level of precision are described. (authors)

  15. Compact optical technique for streak camera calibration

    International Nuclear Information System (INIS)

    To produce accurate data from optical streak cameras requires accurate temporal calibration sources. We have reproduced an older technology for generating optical timing marks that had been lost due to component availability. Many improvements have been made which allow the modern units to service a much larger need. Optical calibrators are now available that produce optical pulse trains of 780 nm wavelength light at frequencies ranging from 0.1 to 10 GHz, with individual pulse widths of approximately 25 ps full width half maximum. Future plans include the development of single units that produce multiple frequencies to cover a wide temporal range, and that are fully controllable via an RS232 interface

  16. Short on camera geometry and camera calibration

    OpenAIRE

    Magnusson, Maria

    2010-01-01

    We will present the basic theory for the camera geometry. Our goal is camera calibration and the tools necessary for this. We start with homogeneous matrices that can be used to describe geometric transformations in a simple manner. Then we consider the pinhole camera model, the simplified camera model that we will show how to calibrate. A camera matrix describes the mapping from the 3D world to a camera image. The camera matrix can be determined through a number of corresponding points measu...

  17. Calibrating Images from the MINERVA Cameras

    Science.gov (United States)

    Mercedes Colón, Ana

    2016-01-01

    The MINiature Exoplanet Radial Velocity Array (MINERVA) consists of an array of robotic telescopes located on Mount Hopkins, Arizona with the purpose of performing transit photometry and spectroscopy to find Earth-like planets around Sun-like stars. In order to make photometric observations, it is necessary to perform calibrations on the CCD cameras of the telescopes to take into account possible instrument error on the data. In this project, we developed a pipeline that takes optical images, calibrates them using sky flats, darks, and biases to generate a transit light curve.

  18. Calibration of radiation protection monitoring equipment

    International Nuclear Information System (INIS)

    The facilities and equipment required for calibrating radiation protection monitoring equipment are both complex and expensive. The staff required to operate the equipment and perform the standardization have to be experienced, well trained and technically competent in the use of secondary or primary radiation standards and equipment. Thus an organization cannot commit considerable resources to perform calibration work purely for academic or scientific reasons. In most cases it must be justified by legal requirements pertaining within the country. These legal requirements on measuring devices for radiation protection purposes may be broadly divided into those relating to instruments intended for area monitoring and to those intended for individual monitoring

  19. New volumetric technique for calibrating neutron probes

    International Nuclear Information System (INIS)

    Studies in laboratory conditions were carried out to determine neutron calibration curves. Soils were packer in barrels to attain values of bulk density similar to those in the field. A mariotte reservoir supplies water to the barels in a way that keeps a water table of 10 cm. A water flow is established from this saturated region to the top of the barrel. The water depth is correlated to the numerical integration of the count rate with depth during the process of capillary tise. The coefficient of determination was 0.999. The calibration curve was independent of the initial misture contents of the soil profile. (author)

  20. The TileCal Laser Calibration System

    Science.gov (United States)

    Giangiobbe, Vincent; On Behalf Of The Atlas Tile Calorimeter Group

    TileCal is the central hadronic calorimeter of the ATLAS detector operating at LHC. It is a sampling calorimeter whose active material is made of scintillating plastic tiles. Scintillation light is read by photomultipliers. A Laser system is used to monitor their gain stability. During dedicated calibration runs the Laser system sends via long optical fibers, a monitored amount of light simultaneously to all the ≈10000 photomultipliers of TileCal. This note describes two complementary methods to measure the stability of the photomultipliers gain using the Laser calibration runs. The results of validation tests are presented for both methods and theirrespective performances and limitations are discussed.

  1. Planck 2013 results. V. LFI calibration

    Science.gov (United States)

    Planck Collaboration; Aghanim, N.; Armitage-Caplan, C.; Arnaud, M.; Ashdown, M.; Atrio-Barandela, F.; Aumont, J.; Baccigalupi, C.; Banday, A. J.; Barreiro, R. B.; Battaner, E.; Benabed, K.; Benoît, A.; Benoit-Lévy, A.; Bernard, J.-P.; Bersanelli, M.; Bielewicz, P.; Bobin, J.; Bock, J. J.; Bonaldi, A.; Bonavera, L.; Bond, J. R.; Borrill, J.; Bouchet, F. R.; Bridges, M.; Bucher, M.; Burigana, C.; Butler, R. C.; Cappellini, B.; Cardoso, J.-F.; Catalano, A.; Chamballu, A.; Chen, X.; Chiang, L.-Y.; Christensen, P. R.; Church, S.; Colombi, S.; Colombo, L. P. L.; Crill, B. P.; Curto, A.; Cuttaia, F.; Danese, L.; Davies, R. D.; Davis, R. J.; de Bernardis, P.; de Rosa, A.; de Zotti, G.; Delabrouille, J.; Dickinson, C.; Diego, J. M.; Dole, H.; Donzelli, S.; Doré, O.; Douspis, M.; Dupac, X.; Efstathiou, G.; Enßlin, T. A.; Eriksen, H. K.; Finelli, F.; Forni, O.; Frailis, M.; Franceschi, E.; Gaier, T. C.; Galeotta, S.; Ganga, K.; Giard, M.; Giardino, G.; Giraud-Héraud, Y.; Gjerløw, E.; González-Nuevo, J.; Górski, K. M.; Gratton, S.; Gregorio, A.; Gruppuso, A.; Hansen, F. K.; Hanson, D.; Harrison, D.; Henrot-Versillé, S.; Hernández-Monteagudo, C.; Herranz, D.; Hildebrandt, S. R.; Hivon, E.; Hobson, M.; Holmes, W. A.; Hornstrup, A.; Hovest, W.; Huffenberger, K. M.; Jaffe, A. H.; Jaffe, T. R.; Jewell, J.; Jones, W. C.; Juvela, M.; Kangaslahti, P.; Keihänen, E.; Keskitalo, R.; Kisner, T. S.; Knoche, J.; Knox, L.; Kunz, M.; Kurki-Suonio, H.; Lagache, G.; Lähteenmäki, A.; Lamarre, J.-M.; Lasenby, A.; Laureijs, R. J.; Lawrence, C. R.; Leach, S.; Leahy, J. P.; Leonardi, R.; Lesgourgues, J.; Liguori, M.; Lilje, P. B.; Linden-Vørnle, M.; López-Caniego, M.; Lubin, P. M.; Macías-Pérez, J. F.; Maino, D.; Mandolesi, N.; Maris, M.; Marshall, D. J.; Martin, P. G.; Martínez-González, E.; Masi, S.; Massardi, M.; Matarrese, S.; Matthai, F.; Mazzotta, P.; Meinhold, P. R.; Melchiorri, A.; Mendes, L.; Mennella, A.; Migliaccio, M.; Mitra, S.; Moneti, A.; Montier, L.; Morgante, G.; Mortlock, D.; Moss, A.; Munshi, D.; Naselsky, P.; Natoli, P.; Netterfield, C. B.; Nørgaard-Nielsen, H. U.; Novikov, D.; Novikov, I.; O'Dwyer, I. J.; Osborne, S.; Paci, F.; Pagano, L.; Paladini, R.; Paoletti, D.; Partridge, B.; Pasian, F.; Patanchon, G.; Pearson, D.; Peel, M.; Perdereau, O.; Perotto, L.; Perrotta, F.; Pierpaoli, E.; Pietrobon, D.; Plaszczynski, S.; Pointecouteau, E.; Polenta, G.; Ponthieu, N.; Popa, L.; Poutanen, T.; Pratt, G. W.; Prézeau, G.; Prunet, S.; Puget, J.-L.; Rachen, J. P.; Rebolo, R.; Reinecke, M.; Remazeilles, M.; Ricciardi, S.; Riller, T.; Rocha, G.; Rosset, C.; Rossetti, M.; Roudier, G.; Rubiño-Martín, J. A.; Rusholme, B.; Sandri, M.; Santos, D.; Scott, D.; Seiffert, M. D.; Shellard, E. P. S.; Spencer, L. D.; Starck, J.-L.; Stolyarov, V.; Stompor, R.; Sureau, F.; Sutton, D.; Suur-Uski, A.-S.; Sygnet, J.-F.; Tauber, J. A.; Tavagnacco, D.; Terenzi, L.; Toffolatti, L.; Tomasi, M.; Tristram, M.; Tucci, M.; Tuovinen, J.; Türler, M.; Umana, G.; Valenziano, L.; Valiviita, J.; Van Tent, B.; Varis, J.; Vielva, P.; Villa, F.; Vittorio, N.; Wade, L. A.; Wandelt, B. D.; Watson, R.; Wilkinson, A.; Yvon, D.; Zacchei, A.; Zonca, A.

    2014-11-01

    We discuss the methods employed to photometrically calibrate the data acquired by the Low Frequency Instrument on Planck. Our calibration is based on a combination of the orbital dipole plus the solar dipole, caused respectively by the motion of the Planck spacecraft with respect to the Sun and by motion of the solar system with respect to the cosmic microwave background (CMB) rest frame. The latter provides a signal of a few mK with the same spectrum as the CMB anisotropies and is visible throughout the mission. In this data releasewe rely on the characterization of the solar dipole as measured by WMAP. We also present preliminary results (at 44 GHz only) on the study of the Orbital Dipole, which agree with the WMAP value of the solar system speed within our uncertainties. We compute the calibration constant for each radiometer roughly once per hour, in order to keep track of changes in the detectors' gain. Since non-idealities in the optical response of the beams proved to be important, we implemented a fast convolution algorithm which considers the full beam response in estimating the signal generated by the dipole. Moreover, in order to further reduce the impact of residual systematics due to sidelobes, we estimated time variations in the calibration constant of the 30 GHz radiometers (the ones with the largest sidelobes) using the signal of an internal reference load at 4 K instead of the CMB dipole. We have estimated the accuracy of the LFI calibration following two strategies: (1) we have run a set of simulations to assess the impact of statistical errors and systematic effects in the instrument and in the calibration procedure; and (2) we have performed a number of internal consistency checks on the data and on the brightness temperature of Jupiter. Errors in the calibration of this Planck/LFI data release are expected to be about 0.6% at 44 and 70 GHz, and 0.8% at 30 GHz. Both these preliminary results at low and high ℓ are consistent with WMAP results

  2. Time Calibration of the ANTARES Neutrino Telescope

    CERN Document Server

    Aguilar, J A; Albert, A; André, M; Anghinolfi, M; Anton, G; Anvar, S; Ardid, M; Jesus, A C Assis; Astraatmadja, T; Aubert, J J; Auer, R; Baret, B; Basa, S; Bazzotti, M; Bertin, V; Biagi, S; Bigongiari, C; Bou-Cabo, M; Bouwhuis, M C; Brown, A M; Brunner, J; Busto, J; Camarena, F; Capone, A; Carloganu, C; Carminati, G; Carr, J; Cecchini, S; Charvis, Ph; Chiarusi, T; Circella, M; Costantini, H; Cottini, N; Coyle, P; Curtil, C; Decowski, M P; Dekeyser, I; Deschamps, A; Distefano, C; Donzaud, C; Dornic, D; Drouhin, D; Eberl, T; Emanuele, U; Ernenwein, J P; Escoffier, S; Fehr, F; Flaminio, V; Fritsch, U; Fuda, J L; Galata, S; Gay, P; Giacomelli, G; Gómez-González, J P; Graf, K; Guillard, G; Halladjian, G; Hallewell, G; van Haren, H; Heijboer, A J; Hello, Y; Hernández-Rey, J J; Herold, B; Hössl, J; Hsu, C C; de Jong, M; Kadler, M; Kalantar-Nayestanaki, N; Kalekin, O; Kappes, A; Katz, U; Kooijman, P; Kopper, C; Kouchner, A; Kulikovskiy, V; Lahmann, R; Lamare, P; Larosa, G; Lefèvre, D; Lim, G; Presti, D Lo; Loehner, H; Loucatos, S; Lucarelli, F; Mangano, S; Marcelin, M; Margiotta, A; Martinez-Mora, J A; Mazure, A; Montaruli, T; Morganti, M; Moscoso, L; Motz, H; Naumann, C; Neff, M; Palioselitis, D; Pavalas, G E; Payre, P; Petrovic, J; Piattelli, P; Picot-Clemente, N; Picq, C; Popa, V; Pradier, T; Presani, E; Racca, C; Reed, C; Riccobene, G; Richardt, C; Rujoiu, M; Russo, G V; Salesa, F; Sapienza, P; Schöck, F; Schuller, J P; Shanidze, R; Simeone, F; Spies, A; Spurio, M; Steijger, J J M; Stolarczyk, Th; Taiuti, M; Tamburini, C; Tasca, L; Toscano, S; Vallage, B; Van Elewyck, V; Vannoni, G; Vecchi, M; Vernin, P; Wijnker, G; de Wolf, E; Yepes, H; Zaborov, D; Zornoza, J D; Zúñiga, J

    2010-01-01

    The ANTARES deep-sea neutrino telescope comprises a three-dimensional array of photomultipliers to detect the Cherenkov light induced by upgoing relativistic charged particles originating from neutrino interactions in the vicinity of the detector. The large scattering length of light in the deep sea facilitates an angular resolution of a few tenths of a degree for neutrino energies exceeding 10 TeV. In order to achieve this optimal performance, the time calibration procedures should ensure a relative time calibration between the photomultipliers at the level of about 1ns. The methods developed to attain this level of precision are described.

  3. Photometric calibrations for 21st century science

    Energy Technology Data Exchange (ETDEWEB)

    Kent, Stephen; /Fermilab; Kaiser, Mary Elizabeth; /Johns Hopkins U.; Deustua, Susana E.; /Baltimore, Space Telescope Sci.; Smith, J.Allyn; /Austin Peay State U.; Adelman, Saul; /Citadel Military Coll.; Allam, Sahar S.; /Fermilab; Baptista, Brian; /Indiana U.; Bohlin, Ralph C.; /Baltimore, Space Telescope Sci.; Clem, James L.; /Louisiana State U.; Conley, Alex; /Colorado U.; Edelstein, Jerry; /UC, Berkeley, Space Sci. Dept. /NOAO, Tucson

    2009-02-01

    The answers to fundamental science questions in astrophysics, ranging from the history of the expansion of the universe to the sizes of nearby stars, hinge on our ability to make precise measurements of diverse astronomical objects. As our knowledge of the underlying physics of objects improves along with advances in detectors and instrumentation, the limits on our capability to extract science from measurements is set, not by our lack of understanding of the nature of these objects, but rather by the most mundane of all issues: the precision with which we can calibrate observations in physical units. In principle, photometric calibration is a solved problem - laboratory reference standards such as blackbody furnaces achieve precisions well in excess of those needed for astrophysics. In practice, however, transferring the calibration from these laboratory standards to astronomical objects of interest is far from trivial - the transfer must reach outside the atmosphere, extend over 4{pi} steradians of sky, cover a wide range of wavelengths, and span an enormous dynamic range in intensity. Virtually all spectrophotometric observations today are calibrated against one or more stellar reference sources, such as Vega, which are themselves tied back to laboratory standards in a variety of ways. This system's accuracy is not uniform. Selected regions of the electromagnetic spectrum are calibrated extremely well, but discontinuities of a few percent still exist, e.g., between the optical and infrared. Independently, model stellar atmospheres are used to calibrate the spectra of selected white dwarf stars, e.g. the HST system, but the ultimate accuracy of this system should be verified against laboratory sources. Our traditional standard star systems, while sufficient until now, need to be improved and extended in order to serve future astrophysics experiments. This white paper calls for a program to improve upon and expand the current networks of

  4. The Fourth VLBA Calibrator Survey - VCS4

    OpenAIRE

    Petrov, L.; Kovalev, Y. Y.; Fomalont, E.; Gordon, D

    2005-01-01

    This paper presents the fourth extension to the Very Long Baseline Array (VLBA) Calibrator Survey, containing 258 new sources not previously observed with very long baseline interferometry (VLBI). This survey, based on three 24 hour VLBA observing sessions, fills remaining areas on the sky above declination -40 degrees where the calibrator density is less than one source within a 4 degree radius disk at any given direction. The share of these area was reduced from 4.6% to 1.9%. Source positio...

  5. On Calibration of pH Meters

    OpenAIRE

    Da-Ming Zhu; Cheng, K. L.

    2005-01-01

    The calibration of pH meters including the pH glass electrode, ISE electrodes, buffers, and the general background for calibration are reviewed. Understanding of basic concepts of pH, pOH, and electrode mechanism is emphasized. New concepts of pH, pOH, as well as critical examination of activity, and activity coefficients are given. The emergence of new solid state pH electrodes and replacement of the salt bridge with a conducting wire have opened up a new horizon for pH measurements. A pH bu...

  6. Consumption Model Calibration and Related Statistical Problems

    Czech Academy of Sciences Publication Activity Database

    Brabec, Marek; Malý, Marek; Pelikán, Emil; Konár, Ondřej

    Athens: WSEAS Press, 2009 - (Perlovsky, L.; Dionysiou, D.; Kostic, L.; Gonzalez-Concepcion, C.; Jaberg, H.; Mastorakis, N.; Zaharim, A.; Sopian, K.), s. 141-146 ISBN 978-960-474-091-8. [AEBD '09. World Multiconference on Applied Economics, Business and Development . Tenerife (ES), 01.07.2009-03.07.2009] R&D Projects: GA AV ČR 1ET400300513 Institutional research plan: CEZ:AV0Z10300504 Keywords : linear calibration * natural gas consumption modeling * Bayesian approach * statistical model * time-varying calibration * state-space model Subject RIV: JE - Non-nuclear Energetics, Energy Consumption ; Use

  7. Calibration source for electron cyclotron emission measurements

    International Nuclear Information System (INIS)

    A high temperature radiation source has been developed for the absolute calibration of diagnostic instruments for measuring electron cyclotron emission from high temperature plasmas. The source has a radiation area of φ150 mm and can be heated up to 500degC. The measured emissivity of the source is close to unity in the wavelength region between 0.5 and 5 mm. The grating polychromator has been calibrated using the radiation source developed. The obtained temperatures agree with those by the pulse height analysis of soft X-rays and Thomson scattering measurement within 10%. (author)

  8. Medical color displays and their calibration

    Science.gov (United States)

    Fan, Jiahua; Roehrig, Hans; Dallas, W.; Krupinski, Elizabeth

    2009-08-01

    Color displays are increasingly used for medical imaging, replacing the traditional monochrome displays in radiology for multi-modality applications, 3D representation applications, etc. Color displays are also used increasingly because of wide spread application of Tele-Medicine, Tele-Dermatology and Digital Pathology. At this time, there is no concerted effort for calibration procedures for this diverse range of color displays in Telemedicine and in other areas of the medical field. Using a colorimeter to measure the display luminance and chrominance properties as well as some processing software we developed a first attempt to a color calibration protocol for the medical imaging field.

  9. E-book Management with Calibre Software

    OpenAIRE

    N P, Jamsheer

    2013-01-01

    Keeping pace with the technology is a big challenge for modern libraries. Reading habits and requirements of the library users are changing. Libraries have to provide access to the varying forms of information available. In addition to the conventional book lending, users should be given access to e-books. This artcle examines how to manage e-books in a library using calibre e-book management software. Calibre is a highly user friendly personal e-book manager. This free software can be used i...

  10. Cartas del Observatorio N° 2

    OpenAIRE

    Castell, Edmon; Torres Carreño, Guillermo Andrés; Diazgranados, Carlos Nicolás; Mora, Yaneth

    2011-01-01

    El boletín del OAN, "Cartas del Observatorio", es una pieza de comunicación desarrollada entre el Observatorio Astronómico Nacional (OAN) y el Área de Comunicación del SPM. El boletín periódico forma parte de los aportes del SPM al Bicentenario de la Independencia (2010) de Colombia. Este segundo número de Cartas del Observatorio, describe la construcción del edificio, el plan museológico por parte del SPM, una de las piezas importante del OAN, la publicación de la infografía y la partici...

  11. La luz del origen del universo

    Directory of Open Access Journals (Sweden)

    Casas, Alberto

    2015-10-01

    Full Text Available Penzias and Wilson discovered the cosmic microwave background, a luminous echo of the Big Bang, in 1965. The amount of information encoded in this relic of the early universe is truly extraordinary. At present, the most precise data about the origin and structure of the universe comes from the study of this primordial light, the full potential of which has yet to be explored.En 1965 se descubrió el fondo de radiación de microondas, un eco luminoso de la Gran Explosión (Big Bang. La cantidad de información contenida en esta reliquia del universo primitivo es verdaderamente extraordinaria. En la actualidad los datos más precisos sobre el origen y estructura del universo provienen del estudio de esta luz primigenia, que aún no ha sido explorada en todo su potencial.

  12. El equilibrio del consumidor

    OpenAIRE

    León Rodríguez, Mª Dolores; Fernández Pérez, Ana Mª

    2008-01-01

    Obtencin de la recta presupuestaria de un consumidor. Definición y características de las Curvas de indiferencia. Concepto de relación marginal de sustitución. Obtención del equilibrio del consumidor

  13. Summary of KOMPSAT-5 Calibration and Validation

    Science.gov (United States)

    Yang, D.; Jeong, H.; Lee, S.; Kim, B.

    2013-12-01

    Korean Multi-Purpose Satellite 5 (KOMPSAT-5), equipped with high resolution X-band (9.66 GHz) Synthetic Aperture Radar (SAR), is planning to be launched on August 22, 2013. With the satellite's primary mission objective being providing Geographical Information System (GIS), Ocean monitoring and Land management, and Disaster and ENvironment monitoring (GOLDEN), it is expected that its applications for scientific research on geographical processes will be extensive. In order to meet its mission objective, the KOMPSAT-5 will provide three different kinds of SAR imaging modes; High Resolution Mode (1 m resolution, 5 km swath), Standard Mode (3 m resolution, 30 km swath), and Wide Swath Mode (20 m resolution, 100 km swath). The KOMPSAT-5 will be operated in a 550 km sun-synchronous, dawn- dusk orbit with a 28-day ground repeat cycle providing valuable image information on Earth surface day-or-night and even in bad weather condition. After successful launch of the satellite, it will go through Launch and Early Operation (LEOP) and In-Orbit Testing (IOT) period about for 6 months to carry out various tests on satellite bus and payload systems. The satellite bus system will be tested during the first 3 weeks after the launch focusing on the Attitude and Orbit Control Subsystem (AOCS) and Integrated GPS Occultation Receiver (IGOR) calibration. With the completion of bus system test, the SAR payload system will be calibrated during initial In-Flight check period (11 weeks) by the joint effort of Thales Alenia Space Italy (TAS-I) and Korea Aerospace Research Institute (KARI). The pointing and relative calibration will be carried out during this period by analyzing the doppler frequency and antenna beam pattern of reflected microwave signal from selected regions with uniform backscattering coefficients (e.g. Amazon rainforest). A dedicated SAR calibration, called primary calibration, will be allocated at the end of LEOP for 12 weeks to perform thorough calibration activities

  14. Calibration Procedures on Oblique Camera Setups

    Science.gov (United States)

    Kemper, G.; Melykuti, B.; Yu, C.

    2016-06-01

    Beside the creation of virtual animated 3D City models, analysis for homeland security and city planning, the accurately determination of geometric features out of oblique imagery is an important task today. Due to the huge number of single images the reduction of control points force to make use of direct referencing devices. This causes a precise camera-calibration and additional adjustment procedures. This paper aims to show the workflow of the various calibration steps and will present examples of the calibration flight with the final 3D City model. In difference to most other software, the oblique cameras are used not as co-registered sensors in relation to the nadir one, all camera images enter the AT process as single pre-oriented data. This enables a better post calibration in order to detect variations in the single camera calibration and other mechanical effects. The shown sensor (Oblique Imager) is based o 5 Phase One cameras were the nadir one has 80 MPIX equipped with a 50 mm lens while the oblique ones capture images with 50 MPix using 80 mm lenses. The cameras are mounted robust inside a housing to protect this against physical and thermal deformations. The sensor head hosts also an IMU which is connected to a POS AV GNSS Receiver. The sensor is stabilized by a gyro-mount which creates floating Antenna -IMU lever arms. They had to be registered together with the Raw GNSS-IMU Data. The camera calibration procedure was performed based on a special calibration flight with 351 shoots of all 5 cameras and registered the GPS/IMU data. This specific mission was designed in two different altitudes with additional cross lines on each flying heights. The five images from each exposure positions have no overlaps but in the block there are many overlaps resulting in up to 200 measurements per points. On each photo there were in average 110 well distributed measured points which is a satisfying number for the camera calibration. In a first step with the help of

  15. Los reptiles del Delta del Orinoco, Venezuela

    OpenAIRE

    Molina, César; Señaris, josefa; Rivas, Gilson

    2004-01-01

    Como resultado de cinco años (1992-1997) de exploraciones herpetológicas, revisiones bibliográficas y examen de museos nacionales, se presenta un análisis taxonómico, ecológico y biogeográfico preliminar de los reptiles del delta del río Orinoco, Estado Delta Amacuro. Se reconocen 70 especies de reptiles agrupados en tres órdenes, 22 familias y 53 géneros. El orden Squamata es el más diverso, con dominancia de las serpientes de la familias Colubridae y Boidae, y los lagartos de las familias G...

  16. Obtención del tiempo de retardo y del tiempo de concentración para varias cuencas del departamento del Valle del Cauca Getting the lag time and the concentration time for various basin of the Valle del Cauca department

    Directory of Open Access Journals (Sweden)

    Materón M. Hernán

    1989-12-01

    Full Text Available En este trabajo se encontraron expresiones matemáticas, a partir de cuatro características fisiográficas, que permitieron reproducir la respuesta hidrológica de las cuencas del centro y norte del Departamento del Valle del Cauca en términos de los tiempos de retardo y de concentración. Se consideraron diez áreas de drenaje cuyas corrientes son afluentes del río Cauca. Las áreas oscilan entre 11.63 a 317.5 km2. Los resultados obtenidos a partir de métodos calibrados en cuencas extranjeras y con un estudio similar realizado en Antioquia, indicaron diferencias significativas entre estos resultados. Las diferencias fueron más altas en las cuencas de mayor extensión.In this work were found mathematical expressions, upon four phisiografic characteristics that allow the hidrological response of the central and north basins of the Valle del Cauca in terms of the lag and concentration time. Ten areas of drainage which currents are afluents of the Cauca river were considered. The areas go from 11.63 to 317.5 km", The results obtained from the calibrated methods in basis of other countries and with a similar method made in Antioquia, Colombia, indicated significant differences between these results. The differences are higger in basins of more extension.

  17. Las cosas del quehacer

    OpenAIRE

    Sergio C. Staude

    2010-01-01

    El psicoanálisis es una praxis y no una práctica especulativa que solo busca afianzar un saber. Esta praxis encuentra su eje central en la noción de acto que es la invitación —y el sostén— a una experiencia y no a la acumulación de información. Ese acto da cuenta del tiempo de un pasaje, y el artículo destaca dos: el del momento inaugural del psicoanálisis y el de la función del psicoanalista en el devenir del análisis. A ambas las une la responsabilidad ...

  18. ``Calibration-on-the-spot'': How to calibrate an EMCCD camera from its images

    Science.gov (United States)

    Mortensen, Kim I.; Flyvbjerg, Henrik

    In localization-based microscopy, super-resolution is obtained by analyzing isolated diffraction-limited spots imaged, typically, with EMCCD cameras. To compare experiments and calculate localization precision, the photon-to-signal amplification factor is needed but unknown without a calibration of the camera. Here we show how this can be done post festum from just a recorded image. We demonstrate this (i) theoretically, mathematically, (ii) by analyzing images recorded with an EMCCD camera, and (iii) by analyzing simulated EMCCD images for which we know the true values of parameters. In summary, our method of calibration-on-the-spot allows calibration of a camera with unknown settings from old images on file, with no other info needed. Consequently, calibration-on-the-spot also makes future camera calibrations before and after measurements unnecessary, because the calibration is encoded in recorded images during the measurement itself, and can at any later time be decoded with calibration-on-the-spot. This work was supported by the Lundbeck Foundation and the Danish Council for Strategic Research Grant No. 10-092322 (PolyNano).

  19. Signal inference with unknown response: calibration uncertainty renormalized estimator

    CERN Document Server

    Dorn, Sebastian; Greiner, Maksim; Selig, Marco; Böhm, Vanessa

    2014-01-01

    The calibration of a measurement device is crucial for every scientific experiment, where a signal has to be inferred from data. We present CURE, the calibration uncertainty renormalized estimator, to reconstruct a signal and simultaneously the instrument's calibration from the same data without knowing the exact calibration, but its covariance structure. The idea of CURE is starting with an assumed calibration to successively include more and more portions of calibration uncertainty into the signal inference equations and to absorb the resulting corrections into renormalized signal (and calibration) solutions. Thereby, the signal inference and calibration problem turns into solving a single system of ordinary differential equations and can be identified with common resummation techniques used in field theories. We verify CURE by applying it to a simplistic toy example and compare it against existent self-calibration schemes, Wiener filter solutions, and Markov Chain Monte Carlo sampling. We conclude that the...

  20. WIM calibration and data quality management

    Directory of Open Access Journals (Sweden)

    D P G de Wet

    2010-10-01

    Full Text Available Weigh-in-motion (WIM scales are installed on various higher order roads in South Africa to provide traffic loading information for pavement design, strategic planning and law enforcement. Some WIM systems produce anomalies that cannot be satisfactorily explained even by highly experienced professionals. Much of the problem relates to the difficulty in determining the appropriate calibration factors to correct systematic measurement error for WIM systems and the inadequacy of data quality management methods. The author has developed a post-calibration method for WIM data, called the Truck Tractor (TT method, to correct the magnitude of recorded axle loads in retrospect. In addition, it incorporates a series of data quality checks. The TT method is robust, accurate and adequately simple for use on a routine basis for a wide variety of South African WIM systems. The calibration module of the TT method (i.e. the procedure to determine the calibration factor, kTT has been accepted by SANRAL and incorporated into the model it uses to quantify the cost of overloading on toll concessions. Some of the data quality checking concepts are also being considered for further use and threshold values for tests are being refined by SANRAL for this purpose.

  1. Developments in radiocarbon calibration for archaeology

    NARCIS (Netherlands)

    Ramsey, Christopher Bronk; Buck, Caitlin E.; Manning, Sturt W.; Reimer, Paula; van der Plicht, Hans

    2006-01-01

    This update on radiocarbon calibration results from the 19th International Radiocarbon Conference at Oxford in April 2006, and is essential reading for all archaeologists. The way radiocarbon dates and absolute dates relate to each other differs in three periods: back to 12400 cal BR radiocarbon dat

  2. Calibration of the MACHO photometry database

    Energy Technology Data Exchange (ETDEWEB)

    Alcock, A

    1998-10-23

    The MACHO Project is a microlensing survey that monitors the brightnesses of ~60 million stars in the Large Magellanic Cloud (LMC), Small Magellanic Cloud, and Galactic bulge. The database presently contains more photometric measurements than previously recorded in the history of astronomy. We describe the calibration of the MACHO two-color photometry and transformation to the standard Kron-Cousins V and R system. This allows for proper comparison with all other observations on the Kron-Cousins standard system. The highest precision calibrations are for ~9 million stars in the LMC bar. For these stars, independent photometric measurements in field-overlap regions indicate standard deviations δvR = 0.020 mag. Calibrated MACHO photometry data are compared with published photometric sequences and with new Hubble Space Telescope observations. We additionally describe the first application of these calibrated data: the construction of the "efficiency" color-magnitude diagram which will be used to calculate our experimental sensitivity for detecting microlensing in the LMC.

  3. Lunar International Science Coordination/Calibration Targets

    Science.gov (United States)

    Head, J. W.; Issacson, P.; Petro, N.; Runyon, C.; Ohtake, M.; Foing, B.; Grande, M.

    2007-01-01

    A new era of international lunar exploration has begun and will expand over the next four years with data acquired from at least four sophisticated remote sensing missions: KAGUYA (SELENE) [Japan], Chang'E [China], Chandrayaan-l [India], and LRO [United States]. It is recognized that this combined activity at the Moon with modern sophisticated sensors wi II provide unprecedented new information about the Moon and will dramatically improve our understanding of Earth's nearest neighbor. It is anticipated that the blooming of scientific exploration of the Moon by nations involved in space activities will seed and foster peaceful international coordination and cooperation that will benefit all. Summarized here are eight Lunar International Science Coordination/Calibration Targets (L-ISCT) that are intended to a) allow cross-calibration of diverse multi-national instruments and b) provide a focus for training young scientists about a range of lunar science issues. The targets, discussed at several scientific forums, were selected for coordinated science and instrument calibration of orbital data. All instrument teams are encouraged to participate in a coordinated activity of early-release data that will improve calibration and validation of data across independent and diverse instruments.

  4. Power calibration study at the Musashi reactor

    International Nuclear Information System (INIS)

    The Musashi reactor (TRIGA-II,100 kW) initially went critical in January of 1963. The reactor had been used for training, isotope production and medical irradiation for boron neutron capture therapy (1). The initial power calibration was based on the use of a calibrated electrical heater in a calorimetric procedure where the rate of rise of the bulk pit water temperature was measured using 2 kW heaters x 6 pieces. The rate of rise of water temperature was determined to be 0.0474 C/kWh. The reactor was then operated to give the same rate of rise of water temperature. Thus the reactor power was established at the value produced by the electrical heaters. A stirrer for tank water mixing was not used. Recent communications (2)(3) indicated that power calibrations using a stirrer provided a much more uniform mixing, and heating in the reactor tank water which was essential for an accurate calibration. In this paper, the effect of mixing using a stirrer was investigated considering the physical factors such as room temperature, humidity, tank water temperature and it's distributions. The room temperature and humidity around the reactor varies 6-30 and 30-80 %, respectively, depending on four seasons. The heat flow through the surface of the pool was also evaluated because the reactor usually operates without cover on the surface of the pool. (orig.)

  5. Analysis and calibration techniques for superconducting resonators

    Science.gov (United States)

    Cataldo, Giuseppe; Wollack, Edward J.; Barrentine, Emily M.; Brown, Ari D.; Moseley, S. Harvey; U-Yen, Kongpop

    2015-01-01

    A method is proposed and experimentally explored for in-situ calibration of complex transmission data for superconducting microwave resonators. This cryogenic calibration method accounts for the instrumental transmission response between the vector network analyzer reference plane and the device calibration plane. Once calibrated, the observed resonator response is analyzed in detail by two approaches. The first, a phenomenological model based on physically realizable rational functions, enables the extraction of multiple resonance frequencies and widths for coupled resonators without explicit specification of the circuit network. In the second, an ABCD-matrix representation for the distributed transmission line circuit is used to model the observed response from the characteristic impedance and propagation constant. When used in conjunction with electromagnetic simulations, the kinetic inductance fraction can be determined with this method with an accuracy of 2%. Datasets for superconducting microstrip and coplanar-waveguide resonator devices were investigated and a recovery within 1% of the observed complex transmission amplitude was achieved with both analysis approaches. The experimental configuration used in microwave characterization of the devices and self-consistent constraints for the electromagnetic constitutive relations for parameter extraction are also presented.

  6. Teaching Camera Calibration by a Constructivist Methodology

    Science.gov (United States)

    Samper, D.; Santolaria, J.; Pastor, J. J.; Aguilar, J. J.

    2010-01-01

    This article describes the Metrovisionlab simulation software and practical sessions designed to teach the most important machine vision camera calibration aspects in courses for senior undergraduate students. By following a constructivist methodology, having received introductory theoretical classes, students use the Metrovisionlab application to…

  7. SENSITIVITY ANALYSIS IN A LIDARCAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    Angel-Iván García-Moreno

    2015-11-01

    Full Text Available In this paper, variability analysis was performed on the model calibration methodology between a multi-camera system and a LiDAR laser sensor (Light Detection and Ranging. Both sensors are used to digitize urban environments. A practical and complete methodology is presented to predict the error propagation inside the LiDAR-camera calibration. We perform a sensitivity analysis in a local and global way. The local approach analyses the output variance with respect to the input, only one parameter is varied at once. In the global sensitivity approach, all parameters are varied simultaneously and sensitivity indexes are calculated on the total variation range of the input parameters. We quantify the uncertainty behaviour in the intrinsic camera parameters and the relationship between the noisy data of both sensors and their calibration. We calculated the sensitivity indexes by two techniques, Sobol and FAST (Fourier amplitude sensitivity test. Statistics of the sensitivity analysis are displayed for each sensor, the sensitivity ratio in laser-camera calibration data

  8. Solutions to the linear camera calibration problem

    Science.gov (United States)

    Grosky, William I.; Tamburino, Louis A.

    1987-01-01

    The general linear camera calibration problem is formulated and several classification schemes for various subcases of this problem are developed. For each subcase, simple solutions are found that satisfy all necessary constraints. The results improve those already in the literature with respect to simplicity, efficiency, and coverage. However, the classification scheme is not exhaustive.

  9. In situ calibration of acoustic emission sensors

    Czech Academy of Sciences Publication Activity Database

    Kober, Jan; Převorovský, Zdeněk

    Brno: University of Technology, Brno, 2015 - (Mazal, P.), s. 93-97 ISBN 978-80-214-5262-6. [International Workshop NDT in Progress /8./. Praha (CZ), 12.10.2015-14.10.2015] Institutional support: RVO:61388998 Keywords : standardisation * Acoustic Emission (AE) * time reversal * calibration * requency response Subject RIV: BI - Acoustics

  10. 21 CFR 862.1150 - Calibrator.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Calibrator. 862.1150 Section 862.1150 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) MEDICAL DEVICES CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES Clinical Chemistry Test Systems § 862.1150...

  11. A Calibrated Index of Human Development

    Science.gov (United States)

    Lind, Niels

    2010-01-01

    The weightings of the four component indicators of the UNDP's Human Development Index HDI appear to be arbitrary and have not been given justification. This paper develops a variant of the HDI, calculated to reflect peoples' revealed evaluations of education and the productivity of work. The resulting Calibrated human Development Index CDI has a…

  12. Planck 2013 results. V. LFI calibration

    DEFF Research Database (Denmark)

    Planck Collaboration,; Aghanim, N.; Armitage-Caplan, C.;

    2014-01-01

    WMAP value of the Solar System speed to 0.2%. We compute the calibration constant for each radiometer roughly once per hour, in order to keep track of changes in the detectors' gain. Since non-idealities in the optical response of the beams proved to be important, we implemented a fast convolution...

  13. Computerized Techniques for Calibrating Pressure Balances

    Science.gov (United States)

    Simpson, D. I.

    1994-01-01

    Pressure balances are generally calibrated by the cross-floating technique, where the forces acting on two similar devices in hydrostatic equilibrium are compared. It is a skilled and time-consuming process which has not previously lent itself to significant automation; computers have mostly been used only to calculate results after measurements have been taken. The objective of the present work was to develop real-time computerized measurement techniques to ease the calibration task, which would fully integrate into a single package with versatile software for calculating and displaying results. The calibration process is now conducted by studying graphical computer displays which derive their inputs from differential-pressure transducers and capacitance or optical displacement sensors. The mass imbalance between oil-operated pressure balances is calculated by interpolating between changes in piston rate-of-fall. Differential-pressure transducers are used to estimate mass imbalances between gas-operated balances, and a quick in situ method for determining their sensitivity has been developed. The new techniques have been successfully applied to a variety of pressure balance designs and substantial reductions in calibration times have been achieved. Reduced levels of scatter have revealed small systematic differences between gauge and absolute modes of operation.

  14. Tanks for liquids: calibration and errors assessment

    International Nuclear Information System (INIS)

    After a brief reference to some of the problems raised by tanks calibration, two methods, theoretical and experimental are presented, so as to achieve it taking into account measurement errors. The method is applied to the transfer of liquid from one tank to another. Further, a practical example is developed. (author)

  15. Cross-calibration of interferometric SAR data

    DEFF Research Database (Denmark)

    Dall, Jørgen

    2003-01-01

    Generation of digital elevation models from interferometric synthetic aperture radar (SAR) data is a well established technique. Achieving a high geometric fidelity calls for a calibration accounting for inaccurate navigation data and system parameters as well as system imperfections. Fully autom...

  16. VLBA Calibrator Survey: Astrometric and Image Results

    OpenAIRE

    Petrov, Leonid; Grodon, David; Beasley, Anthony; Fomalont, Ed

    2003-01-01

    Positions and maps of 1608 new compact sources were obtained in twelve sessions observed during 1994--2002 at the VLBA network at 8.4/2.3 GHz. These sources are recommended for use as calibrators for phase reference imaging and as geodetic sources for astrometric/geodetic VLBI applications.

  17. Optimal Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, M. H.

    1994-01-01

    Calibration of partial safety factors is considered in general, including classes of structures where no code exists beforehand. The partial safety factors are determined such that the difference between the reliability for the different structures in the class considered and a target reliability...

  18. Bayesian calibration of car-following models

    NARCIS (Netherlands)

    Van Hinsbergen, C.P.IJ.; Van Lint, H.W.C.; Hoogendoorn, S.P.; Van Zuylen, H.J.

    2010-01-01

    Recent research has revealed that there exist large inter-driver differences in car-following behavior such that different car-following models may apply to different drivers. This study applies Bayesian techniques to the calibration of car-following models, where prior distributions on each model p

  19. Retained Gas Sampler Calibration and Simulant Tests

    International Nuclear Information System (INIS)

    This test plan provides a method for calibration of the retained gas sampler (RGS) for ammonia gas analysis. Simulant solutions of ammonium hydroxide at known concentrations will be diluted with isotopically labeled 0.04 M ammonium hydroxide solution. Sea sand solids will also be mixed with ammonium hydroxide solution and diluent to determine the accuracy of the system for ammonia gas analysis

  20. Calibrating Gyrochronology using Kepler Asteroseismic targets

    CERN Document Server

    Angus, Ruth; Foreman-Mackey, Daniel; McQuillan, Amy

    2015-01-01

    Among the available methods for dating stars, gyrochronology is a powerful one because it requires knowledge of only the star's mass and rotation period. Gyrochronology relations have previously been calibrated using young clusters, with the Sun providing the only age dependence, and are therefore poorly calibrated at late ages. We used rotation period measurements of 310 Kepler stars with asteroseismic ages, 50 stars from the Hyades and Coma Berenices clusters and 6 field stars (including the Sun) with precise age measurements to calibrate the gyrochronology relation, whilst fully accounting for measurement uncertainties in all observable quantities. We calibrated a relation of the form $P=A^n\\times(B-V-c)^b$, where $P$ is rotation period in days, $A$ is age in Myr, $B$ and $V$ are magnitudes and $a$, $b$ and $n$ are the free parameters of our model. We found $a = 0.40^{+0.3}_{-0.05}$, $b = 0.31^{+0.05}_{-0.02}$ and $n = 0.55^{+0.02}_{-0.09}$. Markov Chain Monte Carlo methods were used to explore the posteri...

  1. The MINOS light-injection calibration system

    International Nuclear Information System (INIS)

    A description is given of the light-injection calibration system that has been developed for the MINOS long-baseline neutrino oscillation experiment. The system is based upon pulsed blue LEDs monitored by PIN photodiodes. It is designed to measure non-linearities in the PMT gain curves, as well as monitoring any drifts in PMT gain, at the 1% level

  2. Direct calibration of PICKY-designed microarrays

    Directory of Open Access Journals (Sweden)

    Ronald Pamela C

    2009-10-01

    Full Text Available Abstract Background Few microarrays have been quantitatively calibrated to identify optimal hybridization conditions because it is difficult to precisely determine the hybridization characteristics of a microarray using biologically variable cDNA samples. Results Using synthesized samples with known concentrations of specific oligonucleotides, a series of microarray experiments was conducted to evaluate microarrays designed by PICKY, an oligo microarray design software tool, and to test a direct microarray calibration method based on the PICKY-predicted, thermodynamically closest nontarget information. The complete set of microarray experiment results is archived in the GEO database with series accession number GSE14717. Additional data files and Perl programs described in this paper can be obtained from the website http://www.complex.iastate.edu under the PICKY Download area. Conclusion PICKY-designed microarray probes are highly reliable over a wide range of hybridization temperatures and sample concentrations. The microarray calibration method reported here allows researchers to experimentally optimize their hybridization conditions. Because this method is straightforward, uses existing microarrays and relatively inexpensive synthesized samples, it can be used by any lab that uses microarrays designed by PICKY. In addition, other microarrays can be reanalyzed by PICKY to obtain the thermodynamically closest nontarget information for calibration.

  3. Beam Calibration of Radio Telescopes with Drones

    Science.gov (United States)

    Chang, Chihway; Monstein, Christian; Refregier, Alexandre; Amara, Adam; Glauser, Adrian; Casura, Sarah

    2015-11-01

    We present a multifrequency far-field beam map for the 5-m dish telescope at the Bleien Observatory measured using a commercially available drone. We describe the hexacopter drone used in this experiment, the design of the flight pattern, and the data analysis scheme. This is the first application of this calibration method to a single-dish radio telescope in the far-field. The high signal-to-noise ratio data allows us to characterize the beam pattern with high accuracy out to at least the fourth side-lobe. The resulting two-dimensional beam pattern is compared with that derived from a more traditional calibration approach using an astronomical calibration source. We discuss the advantages of this method compared to other beam calibration methods. Our results show that this drone-based technique is very promising for ongoing and future radio experiments, where the knowledge of the beam pattern is key to obtaining high-accuracy cosmological and astronomical measurements.

  4. Space Power Facility Reverberation Chamber Calibration Report

    Science.gov (United States)

    Lewis, Catherine C.; Dolesh, Robert J.; Garrett, Michael J.

    2014-01-01

    This document describes the process and results of calibrating the Space Environmental Test EMI Test facility at NASA Plum Brook Space Power Facility according to the specifications of IEC61000-4-21 for susceptibility testing from 100 MHz to 40 GHz. The chamber passed the field uniformity test, in both the empty and loaded conditions, making it the world's largest Reverberation Chamber.

  5. Spectral emissivity of surface blackbody calibrators

    DEFF Research Database (Denmark)

    Clausen, Sønnik

    2007-01-01

    The normal spectral emissivity of commercial infrared calibrators is compared with measurements of anodized aluminum samples and grooved aluminum surfaces coated with Pyromark. Measurements performed by FTIR spectroscopy in the wavelength interval from 2 to 20 mu m and at temperatures between 5 and...

  6. Retained Gas Sampler Calibration and Simulant Tests

    Energy Technology Data Exchange (ETDEWEB)

    CRAWFORD, B.A.

    2000-01-05

    This test plan provides a method for calibration of the retained gas sampler (RGS) for ammonia gas analysis. Simulant solutions of ammonium hydroxide at known concentrations will be diluted with isotopically labeled 0.04 M ammonium hydroxide solution. Sea sand solids will also be mixed with ammonium hydroxide solution and diluent to determine the accuracy of the system for ammonia gas analysis.

  7. The LOFAR long baseline snapshot calibrator survey

    CERN Document Server

    Moldón, J; Wucknitz, O; Jackson, N; Drabent, A; Carozzi, T; Conway, J; Kapińska, A D; McKean, P; Morabito, L; Varenius, E; Zarka, P; Anderson, J; Asgekar, A; Avruch, I M; Bell, M E; Bentum, M J; Bernardi, G; Best, P; Bîrzan, L; Bregman, J; Breitling, F; Broderick, J W; Brüggen, M; Butcher, H R; Carbone, D; Ciardi, B; de Gasperin, F; de Geus, E; Duscha, S; Eislöffel, J; Engels, D; Falcke, H; Fallows, R A; Fender, R; Ferrari, C; Frieswijk, W; Garrett, M A; Grießmeier, J; Gunst, A W; Hamaker, J P; Hassall, T E; Heald, G; Hoeft, M; Juette, E; Karastergiou, A; Kondratiev, V I; Kramer, M; Kuniyoshi, M; Kuper, G; Maat, P; Mann, G; Markoff, S; McFadden, R; McKay-Bukowski, D; Morganti, R; Munk, H; Norden, M J; Offringa, A R; Orru, E; Paas, H; Pandey-Pommier, M; Pizzo, R; Polatidis, A G; Reich, W; Röttgering, H; Rowlinson, A; Scaife, A M M; Schwarz, D; Sluman, J; Smirnov, O; Stappers, B W; Steinmetz, M; Tagger, M; Tang, Y; Tasse, C; Thoudam, S; Toribio, M C; Vermeulen, R; Vocks, C; van Weeren, R J; White, S; Wise, M W; Yatawatta, S; Zensus, A

    2014-01-01

    Aims. An efficient means of locating calibrator sources for International LOFAR is developed and used to determine the average density of usable calibrator sources on the sky for subarcsecond observations at 140 MHz. Methods. We used the multi-beaming capability of LOFAR to conduct a fast and computationally inexpensive survey with the full International LOFAR array. Sources were pre-selected on the basis of 325 MHz arcminute-scale flux density using existing catalogues. By observing 30 different sources in each of the 12 sets of pointings per hour, we were able to inspect 630 sources in two hours to determine if they possess a sufficiently bright compact component to be usable as LOFAR delay calibrators. Results. Over 40% of the observed sources are detected on multiple baselines between international stations and 86 are classified as satisfactory calibrators. We show that a flat low-frequency spectrum (from 74 to 325 MHz) is the best predictor of compactness at 140 MHz. We extrapolate from our sample to sho...

  8. Planck 2015 results. V. LFI calibration

    CERN Document Server

    Ade, P A R; Ashdown, M; Aumont, J; Baccigalupi, C; Banday, A J; Barreiro, R B; Bartolo, N; Battaglia, P; Battaner, E; Benabed, K; Benoît, A; Benoit-Lévy, A; Bernard, J -P; Bersanelli, M; Bielewicz, P; Bock, J J; Bonaldi, A; Bonavera, L; Bond, J R; Borrill, J; Bouchet, F R; Bucher, M; Burigana, C; Butler, R C; Calabrese, E; Cardoso, J -F; Catalano, A; Chamballu, A; Christensen, P R; Colombi, S; Colombo, L P L; Crill, B P; Curto, A; Cuttaia, F; Danese, L; Davies, R D; Davis, R J; de Bernardis, P; de Rosa, A; de Zotti, G; Delabrouille, J; Dickinson, C; Diego, J M; Dole, H; Donzelli, S; Doré, O; Douspis, M; Ducout, A; Dupac, X; Efstathiou, G; Elsner, F; Enßlin, T A; Eriksen, H K; Fergusson, J; Finelli, F; Forni, O; Frailis, M; Franceschi, E; Frejsel, A; Galeotta, S; Galli, S; Ganga, K; Giard, M; Giraud-Héraud, Y; Gjerløw, E; González-Nuevo, J; Górski, K M; Gratton, S; Gregorio, A; Gruppuso, A; Hansen, F K; Hanson, D; Harrison, D L; Henrot-Versillé, S; Herranz, D; Hildebrandt, S R; Hivon, E; Hobson, M; Holmes, W A; Hornstrup, A; Hovest, W; Huffenberger, K M; Hurier, G; Jaffe, A H; Jaffe, T R; Juvela, M; Keihänen, E; Keskitalo, R; Kisner, T S; Knoche, J; Krachmalnicoff, N; Kunz, M; Kurki-Suonio, H; Lagache, G; Lähteenmäki, A; Lamarre, J -M; Lasenby, A; Lattanzi, M; Lawrence, C R; Leahy, J P; Leonardi, R; Lesgourgues, J; Levrier, F; Liguori, M; Lilje, P B; Linden-Vørnle, M; López-Caniego, M; Lubin, P M; Macías-Pérez, J F; Maggio, G; Maino, D; Mandolesi, N; Mangilli, A; Maris, M; Martin, P G; Martínez-González, E; Masi, S; Matarrese, S; McGehee, P; Meinhold, P R; Melchiorri, A; Mendes, L; Mennella, A; Migliaccio, M; Mitra, S; Montier, L; Morgante, G; Mortlock, D; Moss, A; Munshi, D; Murphy, J A; Naselsky, P; Nati, F; Natoli, P; Netterfield, C B; Nørgaard-Nielsen, H U; Novikov, D; Novikov, I; Paci, F; Pagano, L; Pajot, F; Paoletti, D; Partridge, B; Pasian, F; Patanchon, G; Pearson, T J; Peel, M; Perdereau, O; Perotto, L; Perrotta, F; Pettorino, V; Piacentini, F; Pierpaoli, E; Pietrobon, D; Pointecouteau, E; Polenta, G; Pratt, G W; Prézeau, G; Prunet, S; Puget, J -L; Rachen, J P; Rebolo, R; Reinecke, M; Remazeilles, M; Renzi, A; Rocha, G; Romelli, E; Rosset, C; Rossetti, M; Roudier, G; Rubiño-Martín, J A; Rusholme, B; Sandri, M; Santos, D; Savelainen, M; Scott, D; Seiffert, M D; Shellard, E P S; Spencer, L D; Stolyarov, V; Sutton, D; Suur-Uski, A -S; Sygnet, J -F; Tauber, J A; Tavagnacco, D; Terenzi, L; Toffolatti, L; Tomasi, M; Tristram, M; Tucci, M; Tuovinen, J; Türler, M; Umana, G; Valenziano, L; Valiviita, J; Van Tent, B; Vassallo, T; Vielva, P; Villa, F; Wade, L A; Wandelt, B D; Watson, R; Wehus, I K; Wilkinson, A; Yvon, D; Zacchei, A; Zonca, A

    2015-01-01

    We present a description of the pipeline used to calibrate the Planck Low Frequency Instrument (LFI) timelines into thermodynamic temperatures for the Planck 2015 data release, covering 4 years of uninterrupted operations. As in the 2013 data release, our calibrator is provided by the spin-synchronous modulation of the CMB dipole, exploiting both the orbital and solar components. Our 2015 LFI analysis provides an independent Solar dipole estimate in excellent agreement with that of HFI and within $1\\sigma$ (0.3 % in amplitude) of the WMAP value. This 0.3 % shift in the peak-to-peak dipole temperature from WMAP and a global overhaul of the iterative calibration code increases the overall level of the LFI maps by 0.45 % (30 GHz), 0.64 % (44 GHz), and 0.82 % (70 GHz) in temperature with respect to the 2013 Planck data release, thus reducing the discrepancy with the power spectrum measured by WMAP. We estimate that the LFI calibration uncertainty is at the level of 0.20 % for the 70 GHz map, 0.26 % for the 44 GHz...

  9. Beam calibration of radio telescopes with drones

    CERN Document Server

    Chang, Chihway; Refregier, Alexandre; Amara, Adam; Glauser, Adrian; Casura, Sarah

    2015-01-01

    We present a multi-frequency far-field beam map for the 5m dish telescope at the Bleien Observatory measured using a commercially available drone. We describe the hexacopter drone used in this experiment, the design of the flight pattern, and the data analysis scheme. This is the first application of this calibration method to a single dish radio telescope in the far-field. The high signal-to-noise data allows us to characterise the beam pattern with high accuracy out to at least the 4th side-lobe. The resulting 2D beam pattern is compared with that derived from a more traditional calibration approach using an astronomical calibration source. We discuss the advantages of this method compared to other beam calibration methods. Our results show that this drone-based technique is very promising for ongoing and future radio experiments, where the knowledge of the beam pattern is key to obtaining high-accuracy cosmological and astronomical measurements.

  10. Self-Calibration of a Moving Camera by Pre-Calibration

    OpenAIRE

    Sturm, Peter

    1996-01-01

    We consider the problem of self-calibrating a moving camera which is equipped with a zoom lens. This consists essentially in estimating the 5 intrinsic parameters of the pinhole camera model. However, these parameters are not independent. Thus, we propose to do a pre-calibration of the camera, with the aim to model the interdependence of the intrinsic parameters. We show that self-calibration then comes down to the estimation of only 1 intrinsic parameter. We propose a method which exploits t...

  11. Robust calibration of a global aerosol model

    Science.gov (United States)

    Lee, L.; Carslaw, K. S.; Pringle, K. J.; Reddington, C.

    2013-12-01

    Comparison of models and observations is vital for evaluating how well computer models can simulate real world processes. However, many current methods are lacking in their assessment of the model uncertainty, which introduces questions regarding the robustness of the observationally constrained model. In most cases, models are evaluated against observations using a single baseline simulation considered to represent the models' best estimate. The model is then improved in some way so that its comparison to observations is improved. Continuous adjustments in such a way may result in a model that compares better to observations but there may be many compensating features which make prediction with the newly calibrated model difficult to justify. There may also be some model outputs whose comparison to observations becomes worse in some regions/seasons as others improve. In such cases calibration cannot be considered robust. We present details of the calibration of a global aerosol model, GLOMAP, in which we consider not just a single model setup but a perturbed physics ensemble with 28 uncertain parameters. We first quantify the uncertainty in various model outputs (CCN, CN) for the year 2008 and use statistical emulation to identify which of the 28 parameters contribute most to this uncertainty. We then compare the emulated model simulations in the entire parametric uncertainty space to observations. Regions where the entire ensemble lies outside the error of the observations indicate structural model error or gaps in current knowledge which allows us to target future research areas. Where there is some agreement with the observations we use the information on the sources of the model uncertainty to identify geographical regions in which the important parameters are similar. Identification of regional calibration clusters helps us to use information from observation rich regions to calibrate regions with sparse observations and allow us to make recommendations for

  12. Development of calibration procedures for the electron beam calibration of plane parallel ionization chambers

    International Nuclear Information System (INIS)

    In Finland, plane parallel (pp) ionization chambers have been used more than ten years for absolute dose measurements in electron beams of radiation therapy accelerators, for energies below 15 MeV. Before 1997 all pp chambers were calibrated in a 60Co gamma beam at the Finnish SSDL (STUK) for air kerma. Since 1999 all pp chambers have been calibrated by STUK in accelerator electron beams. The local absolute dose measurements (beam calibrations) at hospitals are verified every second year by independent comparative dose measurements by STUK, carried out by ionization chambers during a site visit. All absolute dose measurements are done in a water phantom. The acceptable conditions of the beam for the calibration are always verified by the measurement of beam profiles and depth doses. In regular site visits at the time of 60Co beam calibrations, discrepancies larger than 3% in comparative dose measurements between STUK and the hospitals were constantly observed with some of the hospitals' chambers, despite the chambers had had recalibrations at a 60Co beam. Based on further studies and preliminary electron beam calibrations of those problematic chambers, it became clear that pp chambers had individual characteristics as for the 60Co beam calibration. Therefore, it was decided to introduce an electron beam calibration as the routine method for the pp chambers. By the help of the electron beam calibrations, discrepancies in the comparative dose measurements between STUK and hospitals generally diminished to less than 1 %. The first electron beam calibrations were made by manual construction and it was not sure if repeatability was good enough. Therefore, a high precision jig for pp chamber calibration was constructed. In the precision jig, all chambers are in fixed positions and the only movement in the calibration process is to slide a sledge to change the reference cylindrical chamber (0,6 cm3) to the pp chamber to be calibrated. The depths of the chambers in water

  13. Photometric Calibration of Consumer Video Cameras

    Science.gov (United States)

    Suggs, Robert; Swift, Wesley, Jr.

    2007-01-01

    Equipment and techniques have been developed to implement a method of photometric calibration of consumer video cameras for imaging of objects that are sufficiently narrow or sufficiently distant to be optically equivalent to point or line sources. Heretofore, it has been difficult to calibrate consumer video cameras, especially in cases of image saturation, because they exhibit nonlinear responses with dynamic ranges much smaller than those of scientific-grade video cameras. The present method not only takes this difficulty in stride but also makes it possible to extend effective dynamic ranges to several powers of ten beyond saturation levels. The method will likely be primarily useful in astronomical photometry. There are also potential commercial applications in medical and industrial imaging of point or line sources in the presence of saturation.This development was prompted by the need to measure brightnesses of debris in amateur video images of the breakup of the Space Shuttle Columbia. The purpose of these measurements is to use the brightness values to estimate relative masses of debris objects. In most of the images, the brightness of the main body of Columbia was found to exceed the dynamic ranges of the cameras. A similar problem arose a few years ago in the analysis of video images of Leonid meteors. The present method is a refined version of the calibration method developed to solve the Leonid calibration problem. In this method, one performs an endto- end calibration of the entire imaging system, including not only the imaging optics and imaging photodetector array but also analog tape recording and playback equipment (if used) and any frame grabber or other analog-to-digital converter (if used). To automatically incorporate the effects of nonlinearity and any other distortions into the calibration, the calibration images are processed in precisely the same manner as are the images of meteors, space-shuttle debris, or other objects that one seeks to

  14. Enhancing Seismic Calibration Research Through Software Automation

    Energy Technology Data Exchange (ETDEWEB)

    Ruppert, S; Dodge, D; Elliott, A; Ganzberger, M; Hauk, T; Matzel, E; Ryall, F

    2004-07-09

    The National Nuclear Security Administration (NNSA) Ground-Based Nuclear Explosion Monitoring Research and Engineering (GNEM R&E) Program has made significant progress enhancing the process of deriving seismic calibrations and performing scientific integration with automation tools. We present an overview of our software automation efforts and framework to address the problematic issues of very large datasets and varied formats utilized during seismic calibration research. The software and scientific automation initiatives directly support the rapid collection of raw and contextual seismic data used in research, provide efficient interfaces for researchers to measure/analyze data, and provide a framework for research dataset integration. The automation also improves the researcher's ability to assemble quality controlled research products for delivery into the NNSA Knowledge Base (KB). The software and scientific automation tasks provide the robust foundation upon which synergistic and efficient development of, GNEM R&E Program, seismic calibration research may be built. The task of constructing many seismic calibration products is labor intensive and complex, hence expensive. However, aspects of calibration product construction are susceptible to automation and future economies. We are applying software and scientific automation to problems within two distinct phases or 'tiers' of the seismic calibration process. The first tier involves initial collection of waveform and parameter (bulletin) data that comprise the 'raw materials' from which signal travel-time and amplitude correction surfaces are derived and is highly suited for software automation. The second tier in seismic research content development activities include development of correction surfaces and other calibrations. This second tier is less susceptible to complete automation, as these activities require the judgment of scientists skilled in the interpretation of often highly

  15. Calibration of the MEDUSA neutron spectrometer (abstract)

    International Nuclear Information System (INIS)

    The MEDUSA array is a multielement, scintillator-based neutron time-of-flight spectrometer designed primarily to measure primary and secondary neutron production from indirect drive DD and DT capsule implosions at the Omega Laser in Rochester, NY. The array consists of 824 identical scintillator-photomultiplier tube detectors coupled to analog signal discriminators and high resolution, multihit time-to-digital converters, and is located 19.4 m from the center of the Omega target chamber. It is possible to accurately measure the neutron energy spectrum by simply measuring an adequate sample of neutron flight times to the array (the burn time width is negligible). However it is essential to understand the response of the array detectors to the fusion neutrons before an energy spectrum can be deduced from the data. This array response function is generally given in terms of a calibration constant that relates the expected number of detector hits in the array to the number of source neutrons. The calibration constant is a function of the individual detector gains, the thresholds of the discriminators, and the amount of neutron attenuating material between the array and the target. After gain matching the detectors, a calibration constant can be generated by comparing the array response against a known yield of neutrons (this requires dozens of implosions) or from a first principles measurement of the individual detector efficiencies. In this article, we report on the results of both calibrations of the MEDUSA array. In particular, we will focus on the issues and errors associated with the very different measurements required and discuss a new technique being considered for rapid in situ future calibrations

  16. Calibrating ensemble reliability whilst preserving spatial structure

    Directory of Open Access Journals (Sweden)

    Jonathan Flowerdew

    2014-03-01

    Full Text Available Ensemble forecasts aim to improve decision-making by predicting a set of possible outcomes. Ideally, these would provide probabilities which are both sharp and reliable. In practice, the models, data assimilation and ensemble perturbation systems are all imperfect, leading to deficiencies in the predicted probabilities. This paper presents an ensemble post-processing scheme which directly targets local reliability, calibrating both climatology and ensemble dispersion in one coherent operation. It makes minimal assumptions about the underlying statistical distributions, aiming to extract as much information as possible from the original dynamic forecasts and support statistically awkward variables such as precipitation. The output is a set of ensemble members preserving the spatial, temporal and inter-variable structure from the raw forecasts, which should be beneficial to downstream applications such as hydrological models. The calibration is tested on three leading 15-d ensemble systems, and their aggregation into a simple multimodel ensemble. Results are presented for 12 h, 1° scale over Europe for a range of surface variables, including precipitation. The scheme is very effective at removing unreliability from the raw forecasts, whilst generally preserving or improving statistical resolution. In most cases, these benefits extend to the rarest events at each location within the 2-yr verification period. The reliability and resolution are generally equivalent or superior to those achieved using a Local Quantile-Quantile Transform, an established calibration method which generalises bias correction. The value of preserving spatial structure is demonstrated by the fact that 3×3 averages derived from grid-scale precipitation calibration perform almost as well as direct calibration at 3×3 scale, and much better than a similar test neglecting the spatial relationships. Some remaining issues are discussed regarding the finite size of the output

  17. A calibration free vector network analyzer

    Science.gov (United States)

    Kothari, Arpit

    Recently, two novel single-port, phase-shifter based vector network analyzer (VNA) systems were developed and tested at X-band (8.2--12.4 GHz) and Ka-band (26.4--40 GHz), respectively. These systems operate based on electronically moving the standing wave pattern, set up in a waveguide, over a Schottky detector and sample the standing wave voltage for several phase shift values. Once this system is fully characterized, all parameters in the system become known and hence theoretically, no other correction (or calibration) should be required to obtain the reflection coefficient, (Gamma), of an unknown load. This makes this type of VNA "calibration free" which is a significant advantage over other types of VNAs. To this end, a VNA system, based on this design methodology, was developed at X-band using several design improvements (compared to the previous designs) with the aim of demonstrating this "calibration-free" feature. It was found that when a commercial VNA (HP8510C) is used as the source and the detector, the system works as expected. However, when a detector is used (Schottky diode, log detector, etc.), obtaining correct Gamma still requires the customary three-load calibration. With the aim of exploring the cause, a detailed sensitivity analysis of prominent error sources was performed. Extensive measurements were done with different detection techniques including use of a spectrum analyzer as power detector. The system was tested even for electromagnetic compatibility (EMC) which may have contributed to this issue. Although desired results could not be obtained using the proposed standing-wave-power measuring devices like the Schottky diode but the principle of "calibration-free VNA" was shown to be true.

  18. SWIR calibration of Spectralon reflectance factor

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.; Cooksey, Catherine; Ding, Leibo; Thome, Kurtis J.

    2011-11-01

    Satellite instruments operating in the reflective solar wavelength region require accurate and precise determination of the Bidirectional Reflectance Factor (BRF) of laboratory-based diffusers used in their pre-flight and on-orbit radiometric calibrations. BRF measurements are required throughout the reflected-solar spectrum from the ultraviolet through the shortwave infrared. Spectralon diffusers are commonly used as a reflectance standard for bidirectional and hemispherical geometries. The Diffuser Calibration Laboratory (DCaL) at NASA's Goddard Space Flight Center is a secondary calibration facility with reflectance measurements traceable to those made by the Spectral Tri-function Automated Reference Reflectometer (STARR) facility at the National Institute of Standards and Technology (NIST). For more than two decades, the DCaL has provided numerous NASA projects with BRF data in the ultraviolet (UV), visible (VIS) and the Near InfraRed (NIR) spectral regions. Presented in this paper are measurements of BRF from 1475 nm to 1625 nm obtained using an indium gallium arsenide detector and a tunable coherent light source. The sample was a 50.8 mm (2 in) diameter, 99% white Spectralon target. The BRF results are discussed and compared to empirically generated data from a model based on NIST certified values of 6°directional-hemispherical spectral reflectance factors from 900 nm to 2500 nm. Employing a new NIST capability for measuring bidirectional reflectance using a cooled, extended InGaAs detector, BRF calibration measurements of the same sample were also made using NIST's STARR from 1475 nm to 1625 nm at an incident angle of 0° and at viewing angle of 45°. The total combined uncertainty for BRF in this ShortWave Infrared (SWIR) range is less than 1%. This measurement capability will evolve into a BRF calibration service in SWIR region in support of NASA remote sensing missions.

  19. Elementos del sistema electoral del Congreso Mexicano

    Directory of Open Access Journals (Sweden)

    Aldo Medina García

    2013-08-01

    Full Text Available El presente artículo es un estudio de los componentes del sistema electoral, diseñado por el legislador mexicano para integrar al parlamento federal. Estudio desarrollado a sabiendas de la importancia que reviste el sistema electoral como un medio que permite a los ciudadanos convertir su voto expresado en las urnas en escaños a favor de determinado candidato o partido. En dicho análisis tomamos en cuenta la evolución constitucional del sistema electoral mexicano, la participación de las fuerzas minoritarias en la integración del Senado y de la Cámara de Diputados, la división que se hace del territorio en circunscripciones electorales, la barrera electoral que se fija a los partidos políticos para acceder a la representación en el Parlamento, la fórmula de asignación que se aplica a la votación obtenida, para finalmente concluir que el sistema electoral mexicano es esencialmente mayoritario aun cuando se vista de proporcional.

  20. OBJETIVO DEL DESARROLLO DEL MILENIO 7

    Directory of Open Access Journals (Sweden)

    Margarita M. Pérez O.

    2013-10-01

    Full Text Available Los Objetivos de Desarrollo del Milenio, también conocidos como los Objetivos del Milenio, constituyeron las metas finales que aprobaron los dirigentes mundiales, 189 jefes de estado y de gobierno en la Cumbre del Milenio de las Naciones Unidas celebrada en Nueva York en el año 2000, en la cual los Estados miembros de la organización reafirmaron su compromiso de luchar por un mundo en el que el desarrollo sostenible y la eliminación de la pobreza tuvieran máxima prioridad1 . En la cumbre se definió la Declaración del Milenio, la cual detalla los objetivos, principios, valores, metas y acciones para erradicar la pobreza Mundial hacia el año 2015, todos ellos ratificados con alto interés y acción estratégica de carácter global, como marco para la medición de los progresos en materia de desarrollo.

  1. La crisi del diritto del lavoro

    Directory of Open Access Journals (Sweden)

    Alessandro Garilli

    2012-01-01

    Full Text Available Riassunto - L’Autore ripercorre in chiave critica l’evoluzione delle politiche legislative in materia di lavoro, alla luce dell’attuale fase di crisi economico finanziaria. Nell’esaminare i principali interventi normativi attuati da inizio secolo, l’Autore mette in evidenza le scelte di metodo e gli elementi di continuità che accomunano i vari provvedimenti, individuandoli nell’affannosa ricerca del punto di equilibrio tra flessibilità e garantismo. La riflessione si chiude con alcuni interrogativi sulle ultime linee di riforma del mercato del lavoro e con l’auspicio che, anche al cospetto delle imprescindibili ragioni dell’economia, il diritto del lavoro possa continuare a svolgere il proprio tradizionale ruolo di disciplina di tutela della persona.   Abstract - The author analyzes, with a critically oriented approach, the development of the employment law, in the light of the current economic and financial crisis. In examining the main regulatory measures implemented since the beginning of the century, the author highlights the methodological choices and the elements of continuity that unite the various measures, focusing them in a hard research of a point of balance between flexibility and security. The discussion concludes with some questions about the last lines of the reform of the labor market and with the hope that, even in the face of compelling reasons of economy, labor law can continue to play its traditional role of governing the protection of person.

  2. LIDAR Velodyne HDL-64E Calibration Using Pattern Planes

    Directory of Open Access Journals (Sweden)

    Gerardo Atanacio-Jiménez

    2011-11-01

    Full Text Available This work describes a method for calibration of the Velodyne HDL‐64E scanning LIDAR system. The principal contribution was expressed by a pattern calibration signature, the mathematical model and the numerical algorithm for computing the calibration parameters of the LIDAR. In this calibration pattern the main objective is to minimize systematic errors due to geometric calibration factor. It describes an algorithm for solution of the intrinsic and extrinsic parameters. Finally, its uncertainty was calculated from the standard deviation of calibration result errors.

  3. BUNDLE ADJUSTMENTS CCD CAMERA CALIBRATION BASED ON COLLINEARITY EQUATION

    Institute of Scientific and Technical Information of China (English)

    Liu Changying; Yu Zhijing; Che Rensheng; Ye Dong; Huang Qingcheng; Yang Dingning

    2004-01-01

    The solid template CCD camera calibration method of bundle adjustments based on collinearity equation is presented considering the characteristics of space large-dimension on-line measurement. In the method, a more comprehensive camera model is adopted which is based on the pinhole model extended with distortions corrections. In the process of calibration, calibration precision is improved by imaging at different locations in the whole measurement space, multi-imaging at the same location and bundle adjustments optimization. The calibration experiment proves that the calibration method is able to fulfill calibration requirement of CCD camera applied to vision measurement.

  4. Laboratory implantation for well type ionization chambers calibration

    International Nuclear Information System (INIS)

    The Radiological Science Laboratory is implanting a service for calibration of well type chambers by IAEA training program. The kerma rate in the air (mu Gy/h) of the linear Cs-137 reference source CDCS-J4 have been determined using a well type chamber Standard Imaging HDR-1000 model, which have been calibrated at Secondary Standard Laboratory Calibration of IAEA, whereas two HDR-1000 Plus chambers were calibrated too, following the same standards. The results were compared with Wisconsin University calibration certification and has demonstrated that well type ionization chamber calibration can be used in brachytherapy for several kinds of radionuclides. (Author)

  5. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  6. A digital calibration method for synthetic aperture radar systems

    Science.gov (United States)

    Larson, Richard W.; Jackson, P. L.; Kasischke, Eric S.

    1988-01-01

    A basic method to calibrate imagery from synthetic aperture radar (SAR) systems is presented. SAR images are calibrated by monitoring all the terms of the radar equation. This procedure includes the use of both external (calibrated reference reflectors) and internal (system-generated calibration signals) sources to monitor the total SAR system transfer function. To illustrate the implementation of the procedure, two calibrated SAR images (X-band, 3.2-cm wavelength) are presented, along with the radar cross-section measurements of specific scenes within each image. The sources of error within the SAR image calibration procedure are identified.

  7. Ruptura del projecte parental i disposici?? dels preembrions

    OpenAIRE

    Farn??s Amor??s, Esther

    2010-01-01

    La tesi doctoral "Ruptura del projecte parental i disposici?? dels preembrions" analitza els efectes del consentiment a les t??cniques de reproducci?? assistida prestat en el marc d'un projecte parental. La tesi ofereix una resposta als conflictes que enfronten exesposos o exconvivents en relaci?? al dest?? dels preembrions sobrants d'un cicle de fecundaci?? "in vitro" iniciat durant la relaci??. Aquests conflictes constitueixen un grup de casos al voltant del qual no hi ha solucions legislat...

  8. Calibrating Historical IR Sensors Using GEO, and AVHRR Infrared Tropical Mean Calibration Models

    Science.gov (United States)

    Scarino, Benjamin; Doelling, David R.; Minnis, Patrick; Gopalan, Arun; Haney, Conor; Bhatt, Rajendra

    2014-01-01

    Long-term, remote-sensing-based climate data records (CDRs) are highly dependent on having consistent, wellcalibrated satellite instrument measurements of the Earth's radiant energy. Therefore, by making historical satellite calibrations consistent with those of today's imagers, the Earth-observing community can benefit from a CDR that spans a minimum of 30 years. Most operational meteorological satellites rely on an onboard blackbody and space looks to provide on-orbit IR calibration, but neither target is traceable to absolute standards. The IR channels can also be affected by ice on the detector window, angle dependency of the scan mirror emissivity, stray-light, and detector-to-detector striping. Being able to quantify and correct such degradations would mean IR data from any satellite imager could contribute to a CDR. Recent efforts have focused on utilizing well-calibrated modern hyper-spectral sensors to intercalibrate concurrent operational IR imagers to a single reference. In order to consistently calibrate both historical and current IR imagers to the same reference, however, another strategy is needed. Large, well-characterized tropical-domain Earth targets have the potential of providing an Earth-view reference accuracy of within 0.5 K. To that effort, NASA Langley is developing an IR tropical mean calibration model in order to calibrate historical Advanced Very High Resolution Radiometer (AVHRR) instruments. Using Meteosat-9 (Met-9) as a reference, empirical models are built based on spatially/temporally binned Met-9 and AVHRR tropical IR brightness temperatures. By demonstrating the stability of the Met-9 tropical models, NOAA-18 AVHRR can be calibrated to Met-9 by matching the AVHRR monthly histogram averages with the Met-9 model. This method is validated with ray-matched AVHRR and Met-9 biasdifference time series. Establishing the validity of this empirical model will allow for the calibration of historical AVHRR sensors to within 0.5 K, and thereby

  9. Calibration curve for germanium spectrometers from solutions calibrated by liquid scintillation counting

    International Nuclear Information System (INIS)

    The beta-gamma emitters ''60Co, ''137 Cs, ''131 I, ''210 Pb y ''129 Iare radionuclides for which the calibration by the CIEMAT/NIST method ispossible with uncertainties less than 1%. We prepared, from standardized solutions of these radionuclides, samples in vials of 20 ml. We obtained the calibration curves, efficiency as a function of energy, for two germanium detectors. (Author) 5 refs

  10. Calibrated griz magnitudes of Tycho stars: All-sky photometric calibration using bright stars

    OpenAIRE

    Ofek, E.O.

    2008-01-01

    Photometric calibration to ~5% level is frequently needed at arbitrary celestial location. However, existing all-sky astronomical catalogs do not reach this accuracy, and time consuming photometric calibration procedures are required. I fitted the Hipparcos B-T and V-T magnitudes, along with the 2MASS J, H, and K magnitudes of Tycho-2-catalog stars with stellar spectral templates. From the best-fit spectral template derived for each star, I calculated its synthetic SDSS griz magnitudes and co...

  11. Sociedad abierta del conocimiento

    OpenAIRE

    Echeverría, Javier

    2007-01-01

    En las sociedades del conocimiento hay dos modelos opuestos de gestión: conocimiento libre y conocimiento propietario. Ambos generan desarrollos tecnológicos e innovaciones, pero el primero favorece la emergencia de sociedades abiertas del conocimiento. Aplicando las propuestas de von Hippel, este artículo muestra que la contraposición entre ambos modelos no sólo afecta a la producción y distribución del conocimiento, también a su almacenamiento y uso. Los usuarios son la principal fuente de ...

  12. Secretos del test extraterrestre

    OpenAIRE

    Hybris, Profesor

    2016-01-01

    Por el parasicólogo y astrólogo Profesor Hybris La prensa mundial ha revelado recientemente los detalles del extraño caso de Asdrúbal Carrasquilla, el camionero chileno que desapareciera misteriosamente en su país, ante el asombro de varios automovilistas que lo vieron desvanecerse en el aire, junto con su camión, en la muy transitada vía entre Las Fraguas y San Venancio, en la provincia chilena de Valverde. El hecho ocurrió a las 16.55 del 15 de diciembre de 1981. A las 16.58 del mismo día, ...

  13. La muerte del mito

    OpenAIRE

    Sánchez-Biosca, Vicente

    1990-01-01

    Habría de ser de nuevo John Ford quien sentenciará la defunción del western en cuanto mito, su imposibilidad de construirse en el vacío de la historia. Y habría de ser en uno de los más lúcidos films de la historia del cine: The Man who shot Liberty Valance (El hombre que mató a Liberty Valance, 1962). Todo en este film posee un reverso: un viaje en ferrocarril abre paso al pueblecito de Shinbone. Pero, en el interior del relato, otro emprendido por el antes pionero y ho...

  14. La escritura del mundo

    OpenAIRE

    Martínez García-Posada, Ángel

    2010-01-01

    El mapa de un arquitecto es un paisaje alternativo, una representación intencionada del medio que se aleja del documento impuesto y apunta la posibilidad de transformarlo, así comienza el proyecto; el sentido inverso, llevar al territorio los signos del plano, remite a la esencia de la arquitectura y su capacidad de alterar el mundo. Nuestras representaciones describen una realidad y a su vez son otra distinta, entre ambos campos se produce un flujo que los liga y los condiciona. La cartograf...

  15. LA ESCRITURA DEL MUNDO

    OpenAIRE

    Ángel Martínez García-Posada

    2010-01-01

    RESUMEN El mapa de un arquitecto es un paisaje alternativo, una representación intencionada del medio que se aleja del documento impuesto y apunta la posibilidad de transformarlo, así comienza el proyecto; el sentido inverso, llevar al territorio los signos del plano, remite a la esencia de la arquitectura y su capacidad de alterar el mundo. Nuestras representaciones describen una realidad y a su vez son otra distinta, entre ambos campos se produce un flujo que los liga y los condiciona. La c...

  16. Calibration of High Frequency MEMS Microphones

    Science.gov (United States)

    Shams, Qamar A.; Humphreys, William M.; Bartram, Scott M.; Zuckewar, Allan J.

    2007-01-01

    Understanding and controlling aircraft noise is one of the major research topics of the NASA Fundamental Aeronautics Program. One of the measurement technologies used to acquire noise data is the microphone directional array (DA). Traditional direction array hardware, consisting of commercially available condenser microphones and preamplifiers can be too expensive and their installation in hard-walled wind tunnel test sections too complicated. An emerging micro-machining technology coupled with the latest cutting edge technologies for smaller and faster systems have opened the way for development of MEMS microphones. The MEMS microphone devices are available in the market but suffer from certain important shortcomings. Based on early experiments with array prototypes, it has been found that both the bandwidth and the sound pressure level dynamic range of the microphones should be increased significantly to improve the performance and flexibility of the overall array. Thus, in collaboration with an outside MEMS design vendor, NASA Langley modified commercially available MEMS microphone as shown in Figure 1 to meet the new requirements. Coupled with the design of the enhanced MEMS microphones was the development of a new calibration method for simultaneously obtaining the sensitivity and phase response of the devices over their entire broadband frequency range. Over the years, several methods have been used for microphone calibration. Some of the common methods of microphone calibration are Coupler (Reciprocity, Substitution, and Simultaneous), Pistonphone, Electrostatic actuator, and Free-field calibration (Reciprocity, Substitution, and Simultaneous). Traditionally, electrostatic actuators (EA) have been used to characterize air-condenser microphones for wideband frequency ranges; however, MEMS microphones are not adaptable to the EA method due to their construction and very small diaphragm size. Hence a substitution-based, free-field method was developed to

  17. Compact Radar Transceiver with Included Calibration

    Science.gov (United States)

    McLinden, Matthew; Rincon, Rafael

    2013-01-01

    The Digital Beamforming Synthetic Aperture Radar (DBSAR) is an eight-channel phased array radar system that employs solid-state radar transceivers, a microstrip patch antenna, and a reconfigurable waveform generator and processor unit. The original DBSAR transceiver design utilizes connectorized electronic components that tend to be physically large and heavy. To achieve increased functionality in a smaller volume, PCB (printed circuit board) transceivers were designed to replace the large connectorized transceivers. One of the most challenging problems designing the transceivers in a PCB format was achieving proper performance in the calibration path. For a radar loop-back calibration path, a portion of the transmit signal is coupled out of the antenna feed and fed back into the receiver. This is achieved using passive components for stability and repeatability. Some signal also leaks through the receive path. As these two signal paths are correlated via an unpredictable phase, the leakage through the receive path during transmit must be 30 dB below the calibration path. For DBSAR s design, this requirement called for a 100-dB isolation in the receiver path during transmit. A total of 16 solid-state L-band transceivers on a PCB format were designed. The transceivers include frequency conversion stages, T/R switching, and a calibration path capable of measuring the transmit power-receiver gain product during transmit for pulse-by-pulse calibration or matched filtering. In particular, this calibration path achieves 100-dB isolation between the transmitted signal and the low-noise amplifier through the use of a switching network and a section of physical walls achieving attenuation of radiated leakage. The transceivers were designed in microstrip PCBs with lumped elements and individually packaged components for compactness. Each transceiver was designed on a single PCB with a custom enclosure providing interior walls and compartments to isolate transceiver

  18. Spectral and Radiometric Calibration using Tunable Lasers Project

    Data.gov (United States)

    National Aeronautics and Space Administration —  SIRCUS-based calibration relies on a set of monitoring radiometers and tunable laser sources to provide an absolute radiometric calibration that can approach...

  19. High performance grid for the metric calibration of thermographic cameras

    International Nuclear Information System (INIS)

    The metric calibration of thermographic cameras would make possible the acquisition of geometric data if the principles of photogrammetry are followed. Both the distortion effects introduced in the images by the lens and the perspective effect can be corrected if the calibration parameters are known. This paper presents a calibration grid that allows the automatic metric calibration of thermographic cameras. This calibration grid is made of light and easy-to-find materials to guarantee its portability and handling. The calibration parameters obtained with the presented calibration grid are verified and compared with those obtained with a temperature-based calibration grid through the evaluation of the accuracy and repeatability in the modelling of a standard artefact previously certified

  20. Self-Calibrating Vector Helium Magnetometer (SVHM) Project

    Data.gov (United States)

    National Aeronautics and Space Administration — This Phase 2 SBIR proposal describes the design, fabrication and calibration of a brass-board Self-Calibrating Vector Helium Magnetometer (SVHM). The SVHM...