WorldWideScience

Sample records for c-arm flat detector

  1. Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

    Science.gov (United States)

    Martin, Dennis P; Chapman, Talia; Williamson, Christopher; Tinsley, Brian; Ilyas, Asif M; Wang, Mark L

    2017-11-01

    This study aims to test the hypothesis that: (1) radiation exposure is increased with the intended use of Flat Surface Image Intensifier (FSII) units above the operative surface compared with the traditional below-table configuration; (2) this differential increases in a dose-dependent manner; and (3) radiation exposure varies with body part and proximity to the radiation source. A surgeon mannequin was seated at a radiolucent hand table, positioned for volar distal radius plating. Thermoluminescent dosimeters measured exposure to the eyes, thyroid, chest, hand, and groin, for 1- and 15-minute trials from a mini C-arm FSII unit positioned above and below the operating surface. Background radiation was measured by control dosimeters placed within the operating theater. At 1-minute of exposure, hand and eye dosages were significantly greater with the flat detector positioned above the table. At 15-minutes of exposure, hand radiation dosage exceeded that of all other anatomic sites with the FSII in both positions. Hand exposure was increased in a dose-dependent manner with the flat detector in either position, whereas groin exposure saw a dose-dependent only with the flat detector beneath the operating table. These findings suggest that the surgeon's hands and eyes may incur greater radiation exposure compared with other body parts, during routine mini C-arm FSII utilization in its intended position above the operating table. The clinical impact of these findings remains unclear, and future long-term radiation safety investigation is warranted. Surgeons should take precautions to protect critical body parts, particularly when using FSII technology above the operating with prolonged exposure time.

  2. Evaluation of imaging quality for flat-panel detector based low dose C-arm CT system

    International Nuclear Information System (INIS)

    Seo, Chang-Woo; Cha, Bo Kyung; Jeon, Sungchae; Huh, Young

    2015-01-01

    The image quality associated with the extent of the angle of gantry rotation, the number of projection views, and the dose of X-ray radiation was investigated in flat-panel detector (FPD) based C-arm cone-beam computed tomography (CBCT) system for medical applications. A prototype CBCT system for the projection acquisition used the X-ray tube (A-132, Varian inc.) having rhenium-tungsten molybdenum target and flat panel a-Si X-ray detector (PaxScan 4030CB, Varian inc.) having a 397 x 298 mm active area with 388 μm pixel pitch and 1024 x 768 pixels in 2 by 2 binning mode. The performance comparison of X-ray imaging quality was carried out using the Feldkamp, Davis, and Kress (FDK) reconstruction algorithm between different conditions of projection acquisition. In this work, head-and-dental (75 kVp/20 mA) and chest (90 kVp/25 mA) phantoms were used to evaluate the image quality. The 361 (30 fps x 12 s) projection data during 360 deg. gantry rotation with 1 deg. interval for the 3D reconstruction were acquired. Parke weighting function were applied to handle redundant data and improve the reconstructed image quality in a mobile C-arm system with limited rotation angles. The reconstructed 3D images were investigated for comparison of qualitative image quality in terms of scan protocols (projection views, rotation angles and exposure dose). Furthermore, the performance evaluation in image quality will be investigated regarding X-ray dose and limited projection data for a FPD based mobile C-arm CBCT system. (authors)

  3. Percutaneous sacroplasty with the use of C-arm flat-panel detector CT: technical feasibility and clinical outcome

    International Nuclear Information System (INIS)

    Kang, Sung Eun; Lee, Joon Woo; Kim, Joo Hyung; Kang, Heung Sik; Park, Kun Woo; Yeom, Jin S.

    2011-01-01

    Sacroplasty for sacral insufficiency fractures (SIFs) has been performed mostly under computed tomography (CT) or fluoroscopy guidance. The purposes of this study are to describe technical tips and clinical outcomes of sacroplasty under C-arm flat panel detector CT (C-arm CT) guidance, and to compare the cement distributions shown on C-arm CT with those on multi-detector CT (MDCT). This study consisted of patients who underwent sacroplasty for SIF using C-arm CT from May 2006 to May 2009. Technical success was assessed in terms of cement filling and leakage. Clinical outcome was assessed at short-term (less than 1 month) and long-term (more than 1 month) follow-up using a four-grade patient satisfaction scale: poor, fair, good, and excellent. After sacroplasty, all patients underwent MDCT and three radiologists compared MDCT images with C-arm CT images in consensus, focusing on the cement distribution and cement leakage. Sacroplasties were performed on both sacral alae in all 8 patients (male:female = 2:6, mean age = 76.9, range = 63-82). The technical success rate was 100%. At short-term follow up, 6 patients (87.5%) reported significant improvement. Five patients (62.5%) were available for long-term follow-up and all 5 patients reported a reduced pain and an improved ability to ambulate. Using MDCT as the standard of reference, the cement distribution was visualized equally well by C-arm CT. Sacroplasty under C-arm CT showed excellent technical success and good clinical outcome. There was an excellent correlation between C-arm CT and MDCT in evaluating cement distribution and cement leakage. (orig.)

  4. Percutaneous sacroplasty with the use of C-arm flat-panel detector CT: technical feasibility and clinical outcome

    Energy Technology Data Exchange (ETDEWEB)

    Kang, Sung Eun; Lee, Joon Woo; Kim, Joo Hyung; Kang, Heung Sik [Seoul National University Bundang Hospital, Department of Radiology, Gyeonggi-do (Korea, Republic of); Park, Kun Woo; Yeom, Jin S. [Seoul National University Bundang Hospital, Department of Orthopaedic Surgery, Gyeonggi-do (Korea, Republic of)

    2011-04-15

    Sacroplasty for sacral insufficiency fractures (SIFs) has been performed mostly under computed tomography (CT) or fluoroscopy guidance. The purposes of this study are to describe technical tips and clinical outcomes of sacroplasty under C-arm flat panel detector CT (C-arm CT) guidance, and to compare the cement distributions shown on C-arm CT with those on multi-detector CT (MDCT). This study consisted of patients who underwent sacroplasty for SIF using C-arm CT from May 2006 to May 2009. Technical success was assessed in terms of cement filling and leakage. Clinical outcome was assessed at short-term (less than 1 month) and long-term (more than 1 month) follow-up using a four-grade patient satisfaction scale: poor, fair, good, and excellent. After sacroplasty, all patients underwent MDCT and three radiologists compared MDCT images with C-arm CT images in consensus, focusing on the cement distribution and cement leakage. Sacroplasties were performed on both sacral alae in all 8 patients (male:female = 2:6, mean age = 76.9, range = 63-82). The technical success rate was 100%. At short-term follow up, 6 patients (87.5%) reported significant improvement. Five patients (62.5%) were available for long-term follow-up and all 5 patients reported a reduced pain and an improved ability to ambulate. Using MDCT as the standard of reference, the cement distribution was visualized equally well by C-arm CT. Sacroplasty under C-arm CT showed excellent technical success and good clinical outcome. There was an excellent correlation between C-arm CT and MDCT in evaluating cement distribution and cement leakage. (orig.)

  5. Image quality analysis of vibration effects In C-arm-flat panel X-ray imaging

    NARCIS (Netherlands)

    Snoeren, R.M.; Kroon, J.N.; With, de P.H.N.

    2011-01-01

    The motion of C-arm scanning X-ray systems may result in vibrations of the imaging sub-system. In this paper, we connect C-arm system vibrations to Image Quality (IQ) deterioration for 2D angiography and 3D cone beam X-ray imaging, using large Flat Panel detectors. Vibrations will affect the

  6. C-arm flat detector computed tomography: the technique and its applications in interventional neuro-radiology

    International Nuclear Information System (INIS)

    Kamran, Mudassar; Nagaraja, Sanjoy; Byrne, James V.

    2010-01-01

    Flat detector computed tomography (FDCT) is an imaging tool that generates three-dimensional (3-D) volumes from data obtained during C-arm rotation using CT-like reconstruction algorithms. The technique is relatively new and, at current levels of performance, lags behind conventional CT in terms of image quality. However, the advantage of its availability in the interventional room has prompted neuro-radiologists to identify clinical settings where its role is uniquely beneficial. We performed a search of the online literature databases to identify studies reporting experience with FDCT in interventional neuro-radiology. The studies were systematically reviewed and their findings grouped according to specific clinical situation addressed. FDCT images allow detection of procedural complications, evaluation of low-radiopacity stents and assessment of endosaccular coil packing in intra-cranial aneurysms. Additional roles are 3-D angiography that provides an accurate depiction of vessel morphology with low concentrations of radiographic contrast media and a potential for perfusion imaging due to its dynamic scanning capability. A single scan combining soft tissue and angiographic examinations reduces radiation dose and examination time. Ongoing developments in flat detector technology and reconstruction algorithms are expected to further enhance its performance and increase this range of applications. FDCT images provide useful information in neuro-interventional setting. If current research confirms its potential for assessing cerebral haemodynamics by perfusion scanning, the combination would redefine it as an invaluable tool for interventional neuro-radiology procedures. This facility and its existing capabilities of parenchymal and angiographic imaging would also extend its use to the triage of acute stroke patients. (orig.)

  7. Preliminary performance of image quality for a low-dose C-arm CT system with a flat-panel detector

    Energy Technology Data Exchange (ETDEWEB)

    Kyung Cha, Bo [Advanced Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan (Korea, Republic of); Seo, Chang-Woo [Department of Radiation Convergence Engineering, College of Health Science, Yonsei University, Wonju (Korea, Republic of); Yang, Keedong [Advanced Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan (Korea, Republic of); Jeon, Seongchae, E-mail: sarim@keri.re.kr [Advanced Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan (Korea, Republic of); Huh, Young [Advanced Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan (Korea, Republic of)

    2015-06-01

    Digital flat panel imager (FPI)-based cone-beam computed tomography (CBCT) has been widely used in C-arm imaging for spine surgery and interventional procedures. The system provides real-time fluoroscopy with high spatial resolution and three-dimensional (3D) visualization of anatomical structure without the need for patient transportation in interventional suite. In this work, a prototype CBCT imaging platform with continuous single rotation about the gantry was developed by using a large-area flat-panel detector with amorphous Si-based thin film transistor matrix. The different 2D projection images were acquired during constant gantry velocity for reconstructed images at a tube voltage of 80–120 kVp, and different current (10–50 mA) conditions. Various scan protocols were applied to a chest phantom human by changing the number of projection images and scanning angles. The projections were then reconstructed into a volumetric data of sections by using a 3D reconstruction algorithm (e.g., filtered back projection). The preliminary quantitative X-ray performance of our CBCT system was investigated by using the American Association of Physicists in Medicine CT phantom in terms of spatial resolution, contrast resolution, and CT number linearity for mobile or fixed C-arm based CBCT application with limited rotational geometry. The novel results of the projection data with different scanning angles and angular increments in the orbital gantry platform were acquired and evaluated experimentally.

  8. Preliminary performance of image quality for a low-dose C-arm CT system with a flat-panel detector

    International Nuclear Information System (INIS)

    Kyung Cha, Bo; Seo, Chang-Woo; Yang, Keedong; Jeon, Seongchae; Huh, Young

    2015-01-01

    Digital flat panel imager (FPI)-based cone-beam computed tomography (CBCT) has been widely used in C-arm imaging for spine surgery and interventional procedures. The system provides real-time fluoroscopy with high spatial resolution and three-dimensional (3D) visualization of anatomical structure without the need for patient transportation in interventional suite. In this work, a prototype CBCT imaging platform with continuous single rotation about the gantry was developed by using a large-area flat-panel detector with amorphous Si-based thin film transistor matrix. The different 2D projection images were acquired during constant gantry velocity for reconstructed images at a tube voltage of 80–120 kVp, and different current (10–50 mA) conditions. Various scan protocols were applied to a chest phantom human by changing the number of projection images and scanning angles. The projections were then reconstructed into a volumetric data of sections by using a 3D reconstruction algorithm (e.g., filtered back projection). The preliminary quantitative X-ray performance of our CBCT system was investigated by using the American Association of Physicists in Medicine CT phantom in terms of spatial resolution, contrast resolution, and CT number linearity for mobile or fixed C-arm based CBCT application with limited rotational geometry. The novel results of the projection data with different scanning angles and angular increments in the orbital gantry platform were acquired and evaluated experimentally

  9. Comparison of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease with Flat-Panel Detectors on Mobile C-arm versus Fixed Systems.

    Science.gov (United States)

    Guillou, Marie; Maurel, Blandine; Necib, Hatem; Vent, Pierre-Alexandre; Costargent, Alain; Chaillou, Philippe; Gouëffic, Yann; Kaladji, Adrien

    2018-02-01

    Flat-panel detectors on mobile C-arm (MC-arm) systems are currently challenging fixed C-arm (FC-arm) systems used in hybrid operating rooms. MC-arm systems offer an alternative to FC-arm systems in the endovascular treatment of peripheral arterial disease (PAD) but their efficiency has not been evaluated comparatively. Two series of patients undergoing arteriography with intention to treat were included. Each series consisted of 2 nonrandomized groups: an MC-arm group and an FC-arm group. Series 1 evaluated exposure to the patient (MC-arm, n = 113; FC-arm, n = 206) while series 2 evaluated exposure to patients and also health care personnel (MC-arm, n = 24; FC-arm, n = 76). The primary end points for evaluating exposure were air kerma (AK, in mGy) for patients and effective dose for health care personnel (in μSv). After adjustment for the effect of body mass index (analysis of covariance test), AK was found to be lower in the MC-arm group than in the FC-arm group (124.1 ± 142 vs. 173.3 ± 248.7, P = 0.025). There was no difference between the groups with regard to effective dose recorded for senior surgeons or for operating room nurses. However, a higher effective dose was recorded by the MC-arm group external dosimeter for the trainee resident and for nurse anesthetists. In endovascular treatment of lower limb PAD, use of an FC-arm system is associated with more radiation exposure to the patient than an MC-arm system. However, this type of imaging system does not appear to affect exposure to health care personnel. Copyright © 2017 Elsevier Inc. All rights reserved.

  10. A Comparison of Image Quality and Radiation Exposure Between the Mini C-Arm and the Standard C-Arm.

    Science.gov (United States)

    van Rappard, Juliaan R M; Hummel, Willy A; de Jong, Tijmen; Mouës, Chantal M

    2018-04-01

    The use of intraoperative fluoroscopy has become mandatory in osseous hand surgery. Due to its overall practicality, the mini C-arm has gained popularity among hand surgeons over the standard C-arm. This study compares image quality and radiation exposure for patient and staff between the mini C-arm and the standard C-arm, both with flat panel technology. An observer-based subjective image quality study was performed using a contrast detail (CD) phantom. Five independent observers were asked to determine the smallest circles discernable to them. The results were plotted in a graph, forming a CD curve. From each curve, an image quality figure (IQF) was derived. A lower IQF equates to a better image quality. The patients' entrance skin dose was measured, and to obtain more information about the staff exposure dose, a perspex hand phantom was used. The scatter radiation was measured at various distances and angles relative to a central point on the detector. The IQF was significantly lower for the mini C-arm resulting in a better image quality. The patients' entrance dose was 10 times higher for the mini C-arm as compared with the standard C-arm, and the scatter radiation threefold. Due to its improved image quality and overall practicality, the mini C-arm is recommended for hand surgical procedures. To ensure that the surgeons' radiation exposure is not exceeding the safety limits, monitoring radiation exposure using mini C-arms with flat panel technology during surgery should be done in a future clinical study.

  11. Artifact reduction of different metallic implants in flat detector C-arm CT.

    Science.gov (United States)

    Hung, S-C; Wu, C-C; Lin, C-J; Guo, W-Y; Luo, C-B; Chang, F-C; Chang, C-Y

    2014-07-01

    Flat detector CT has been increasingly used as a follow-up examination after endovascular intervention. Metal artifact reduction has been successfully demonstrated in coil mass cases, but only in a small series. We attempted to objectively and subjectively evaluate the feasibility of metal artifact reduction with various metallic objects and coil lengths. We retrospectively reprocessed the flat detector CT data of 28 patients (15 men, 13 women; mean age, 55.6 years) after they underwent endovascular treatment (20 coiling ± stent placement, 6 liquid embolizers) or shunt drainage (n = 2) between January 2009 and November 2011 by using a metal artifact reduction correction algorithm. We measured CT value ranges and noise by using region-of-interest methods, and 2 experienced neuroradiologists rated the degrees of improved imaging quality and artifact reduction by comparing uncorrected and corrected images. After we applied the metal artifact reduction algorithm, the CT value ranges and the noise were substantially reduced (1815.3 ± 793.7 versus 231.7 ± 95.9 and 319.9 ± 136.6 versus 45.9 ± 14.0; both P metallic objects and various sizes of coil masses. The rater study achieved an overall improvement of imaging quality and artifact reduction (85.7% and 78.6% of cases by 2 raters, respectively), with the greatest improvement in the coiling group, moderate improvement in the liquid embolizers, and the smallest improvement in ventricular shunting (overall agreement, 0.857). The metal artifact reduction algorithm substantially reduced artifacts and improved the objective image quality in every studied case. It also allowed improved diagnostic confidence in most cases. © 2014 by American Journal of Neuroradiology.

  12. Flat-detector computed tomography in diagnostic and interventional neuroradiology

    International Nuclear Information System (INIS)

    Struffert, T.; Doerfler, A.

    2009-01-01

    Originally aimed at improving standard radiography by providing higher absorption efficiency and a wider dynamic range than available with film-screen and phosphor luminescence, radiography flat detector technology is now widely accepted for neuroangiographic imaging. Especially flat-detector computed tomography (FD-CT), which uses rotational C-arm mounted flat-panel detector technology, is capable of volumetric imaging with a high spatial resolution. As ''angiographic CT'' FD-CT may be helpful in many diagnostic and neurointerventional procedures, e.g. intracranial stenting for cerebrovascular stenoses, stent-assisted coil embolization of wide-necked cerebral aneurysms and embolization of arteriovenous malformations. By providing morphologic, CT-like images of the brain within the angiography suite FD-CT allows rapid visualization of periprocedural hemorrhaging and may thus improve rapid complication management without the need of patient transfer. In addition, myelography and postmyelographic FD-CT imaging can be carried out using a single modality. Spinal interventions, such as kyphoplasty or vertebroplasty might also benefit from FD-CT. Imaging of the temporal bone may also develop into an important field of FD-CT. This paper briefly reviews the technical principles of FD technology and the potential applications in diagnostic and interventional neuroradiology. (orig.) [de

  13. C-arm flat detector computed tomography parenchymal blood volume imaging: the nature of parenchymal blood volume parameter and the feasibility of parenchymal blood volume imaging in aneurysmal subarachnoid haemorrhage patients

    Energy Technology Data Exchange (ETDEWEB)

    Kamran, Mudassar; Byrne, James V. [University of Oxford, Nuffield Department of Surgical Sciences, Oxford (United Kingdom)

    2015-09-15

    C-arm flat detector computed tomography (FDCT) parenchymal blood volume (PBV) measurements allow assessment of cerebral haemodynamics in the neurointerventional suite. This paper explores the feasibility of C-arm computed tomography (CT) PBV imaging and the relationship between the C-arm CT PBV and the MR-PWI-derived cerebral blood volume (CBV) and cerebral blood flow (CBF) parameters in aneurysmal subarachnoid haemorrhage (SAH) patients developing delayed cerebral ischemia (DCI). Twenty-six patients with DCI following aneurysmal SAH underwent a research C-arm CT PBV scan using a biplane angiography system and contemporaneous MR-PWI scan as part of a prospective study. Quantitative whole-brain atlas-based volume-of-interest analysis in conjunction with Pearson correlation and Bland-Altman tests was performed to explore the agreement between C-arm CT PBV and MR-derived CBV and CBF measurements. All patients received medical management, while eight patients (31 %) underwent selective intra-arterial chemical angioplasty. Colour-coded C-arm CT PBV maps were 91 % sensitive and 100 % specific in detecting the perfusion abnormalities. C-arm CT rPBV demonstrated good agreement and strong correlation with both MR-rCBV and MR-rCBF measurements; the agreement and correlation were stronger for MR-rCBF relative to MR-rCBV and improved for C-arm CT PBV versus the geometric mean of MR-rCBV and MR-rCBF. Analysis of weighted means showed that the C-arm CT PBV has a preferential blood flow weighting (∼60 % blood flow and ∼40 % blood volume weighting). C-arm CT PBV imaging is feasible in DCI following aneurysmal SAH. PBV is a composite perfusion parameter incorporating both blood flow and blood volume weightings. That PBV has preferential (∼60 %) blood flow weighting is an important finding, which is of clinical significance when interpreting the C-arm CT PBV maps, particularly in the setting of acute brain ischemia. (orig.)

  14. C-arm flat detector computed tomography parenchymal blood volume imaging: the nature of parenchymal blood volume parameter and the feasibility of parenchymal blood volume imaging in aneurysmal subarachnoid haemorrhage patients

    International Nuclear Information System (INIS)

    Kamran, Mudassar; Byrne, James V.

    2015-01-01

    C-arm flat detector computed tomography (FDCT) parenchymal blood volume (PBV) measurements allow assessment of cerebral haemodynamics in the neurointerventional suite. This paper explores the feasibility of C-arm computed tomography (CT) PBV imaging and the relationship between the C-arm CT PBV and the MR-PWI-derived cerebral blood volume (CBV) and cerebral blood flow (CBF) parameters in aneurysmal subarachnoid haemorrhage (SAH) patients developing delayed cerebral ischemia (DCI). Twenty-six patients with DCI following aneurysmal SAH underwent a research C-arm CT PBV scan using a biplane angiography system and contemporaneous MR-PWI scan as part of a prospective study. Quantitative whole-brain atlas-based volume-of-interest analysis in conjunction with Pearson correlation and Bland-Altman tests was performed to explore the agreement between C-arm CT PBV and MR-derived CBV and CBF measurements. All patients received medical management, while eight patients (31 %) underwent selective intra-arterial chemical angioplasty. Colour-coded C-arm CT PBV maps were 91 % sensitive and 100 % specific in detecting the perfusion abnormalities. C-arm CT rPBV demonstrated good agreement and strong correlation with both MR-rCBV and MR-rCBF measurements; the agreement and correlation were stronger for MR-rCBF relative to MR-rCBV and improved for C-arm CT PBV versus the geometric mean of MR-rCBV and MR-rCBF. Analysis of weighted means showed that the C-arm CT PBV has a preferential blood flow weighting (∼60 % blood flow and ∼40 % blood volume weighting). C-arm CT PBV imaging is feasible in DCI following aneurysmal SAH. PBV is a composite perfusion parameter incorporating both blood flow and blood volume weightings. That PBV has preferential (∼60 %) blood flow weighting is an important finding, which is of clinical significance when interpreting the C-arm CT PBV maps, particularly in the setting of acute brain ischemia. (orig.)

  15. C-arm CT for planning and guidance of extrahepatic embolizations

    International Nuclear Information System (INIS)

    Wacker, F.K.; Meissner, O.A.; Meyer, B.C.

    2009-01-01

    Interventional radiological vascular embolizations are complex procedures that require exact imaging of the target region to facilitate safe and effective treatment. The purpose of this paper is to present the technique and feasibility of flat detector C-arm computed tomography (C-arm CT) for control and guidance of extrahepatic abdominal embolization procedures. C-arm CT images can provide important information on both vascular and cross-sectional anatomy of the target region, help in determining therapy endpoints and provide follow-up during and immediately after the abdominal interventions.The cases presented demonstrate that C-arm CT images are beneficial for abdominal embolization procedures and facilitate precise treatment. (orig.) [de

  16. Calibration model of a dual gain flat panel detector for 2D and 3D x-ray imaging

    International Nuclear Information System (INIS)

    Schmidgunst, C.; Ritter, D.; Lang, E.

    2007-01-01

    The continuing research and further development in flat panel detector technology have led to its integration into more and more medical x-ray systems for two-dimensional (2D) and three-dimensional (3D) imaging, such as fixed or mobile C arms. Besides the obvious advantages of flat panel detectors, like the slim design and the resulting optimum accessibility to the patient, their success is primarily a product of the image quality that can be achieved. The benefits in the physical and performance-related features as opposed to conventional image intensifier systems (e.g., distortion-free reproduction of imaging information or almost linear signal response over a large dynamic range) can be fully exploited, however, only if the raw detector images are correctly calibrated and postprocessed. Previous procedures for processing raw data contain idealizations that, in the real world, lead to artifacts or losses in image quality. Thus, for example, temperature dependencies or changes in beam geometry, as can occur with mobile C arm systems, have not been taken into account up to this time. Additionally, adverse characteristics such as image lag or aging effects have to be compensated to attain the best possible image quality. In this article a procedure is presented that takes into account the important dependencies of the individual pixel sensitivity of flat panel detectors used in 2D or 3D imaging and simultaneously minimizes the work required for an extensive recalibration. It is suitable for conventional detectors with only one gain mode as well as for the detectors specially developed for 3D imaging with dual gain read-out technology

  17. Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery

    International Nuclear Information System (INIS)

    Siewerdsen, J.H.; Moseley, D.J.; Burch, S.; Bisland, S.K.; Bogaards, A.; Wilson, B.C.; Jaffray, D.A.

    2005-01-01

    A mobile isocentric C-arm (Siemens PowerMobil) has been modified in our laboratory to include a large area flat-panel detector (in place of the x-ray image intensifier), providing multi-mode fluoroscopy and cone-beam computed tomography (CT) imaging capability. This platform represents a promising technology for minimally invasive, image-guided surgical procedures where precision in the placement of interventional tools with respect to bony and soft-tissue structures is critical. The image quality and performance in surgical guidance was investigated in pre-clinical evaluation in image-guided spinal surgery. The control, acquisition, and reconstruction system are described. The reproducibility of geometric calibration, essential to achieving high three-dimensional (3D) image quality, is tested over extended time scales (7 months) and across a broad range in C-arm angulation (up to 45 deg.), quantifying the effect of improper calibration on spatial resolution, soft-tissue visibility, and image artifacts. Phantom studies were performed to investigate the precision of 3D localization (viz., fiber optic probes within a vertebral body) and effect of lateral projection truncation (limited field of view) on soft-tissue detectability in image reconstructions. Pre-clinical investigation was undertaken in a specific spinal procedure (photodynamic therapy of spinal metastases) in five animal subjects (pigs). In each procedure, placement of fiber optic catheters in two vertebrae (L1 and L2) was guided by fluoroscopy and cone-beam CT. Experience across five procedures is reported, focusing on 3D image quality, the effects of respiratory motion, limited field of view, reconstruction filter, and imaging dose. Overall, the intraoperative cone-beam CT images were sufficient for guidance of needles and catheters with respect to bony anatomy and improved surgical performance and confidence through 3D visualization and verification of transpedicular trajectories and tool placement

  18. Reduced-dose C-arm computed tomography applications at a pediatric institution

    Energy Technology Data Exchange (ETDEWEB)

    Acord, Michael; Shellikeri, Sphoorti; Vatsky, Seth; Srinivasan, Abhay; Krishnamurthy, Ganesh; Keller, Marc S.; Cahill, Anne Marie [The Children' s Hospital of Philadelphia, Department of Radiology, Philadelphia, PA (United States)

    2017-12-15

    Reduced-dose C-arm computed tomography (CT) uses flat-panel detectors to acquire real-time 3-D images in the interventional radiology suite to assist with anatomical localization and procedure planning. To describe dose-reduction techniques for C-arm CT at a pediatric institution and to provide guidance for implementation. We conducted a 5-year retrospective study on procedures using an institution-specific reduced-dose protocol: 5 or 8 s Dyna Rotation, 248/396 projection images/acquisition and 0.1-0.17 μGy/projection dose at the detector with 0.3/0.6/0.9-mm copper (Cu) filtration. We categorized cases by procedure type and average patient age and calculated C-arm CT and total dose area product (DAP). Two hundred twenty-two C-arm CT-guided procedures were performed with a dose-reduction protocol. The most common procedures were temporomandibular and sacroiliac joint injections (48.6%) and sclerotherapy (34.2%). C-arm CT was utilized in cases of difficult percutaneous access in less common applications such as cecostomy and gastrostomy placement, foreign body retrieval and thoracentesis. C-arm CT accounted for between 9.9% and 80.7% of the total procedural DAP. Dose-reducing techniques can preserve image quality for intervention while reducing radiation exposure to the child. This technology has multiple applications within pediatric interventional radiology and can be considered as an adjunctive imaging tool in a variety of procedures, particularly when percutaneous access is challenging despite routine fluoroscopic or ultrasound guidance. (orig.)

  19. Flat-detector computed tomography (FD-CT)

    International Nuclear Information System (INIS)

    Kalender, Willi A.; Kyriakou, Yiannis

    2007-01-01

    Flat-panel detectors or, synonymously, flat detectors (FDs) have been developed for use in radiography and fluoroscopy with the defined goal to replace standard X-ray film, film-screen combinations and image intensifiers by an advanced sensor system. FD technology in comparison to X-ray film and image intensifiers offers higher dynamic range, dose reduction, fast digital readout and the possibility for dynamic acquisitions of image series, yet keeping to a compact design. It appeared logical to employ FD designs also for computed tomography (CT) imaging. Respective efforts date back a few years only, but FD-CT has meanwhile become widely accepted for interventional and intra-operative imaging using C-arm systems. FD-CT provides a very efficient way of combining two-dimensional (2D) radiographic or fluoroscopic and 3D CT imaging. In addition, FD technology made its way into a number of dedicated CT scanner developments, such as scanners for the maxillo-facial region or for micro-CT applications. This review focuses on technical and performance issues of FD technology and its full range of applications for CT imaging. A comparison with standard clinical CT is of primary interest. It reveals that FD-CT provides higher spatial resolution, but encompasses a number of disadvantages, such as lower dose efficiency, smaller field of view and lower temporal resolution. FD-CT is not aimed at challenging standard clinical CT as regards to the typical diagnostic examinations; but it has already proven unique for a number of dedicated CT applications, offering distinct practical advantages, above all the availability of immediate CT imaging in the interventional suite or the operating room. (orig.)

  20. Flat detectors and their clinical applications

    International Nuclear Information System (INIS)

    Spahn, Martin

    2005-01-01

    Diagnostic and interventional flat detector X-ray systems are penetrating the market in all application segments. First introduced in radiography and mammography, they have conquered cardiac and general angiography and are getting increasing attention in fluoroscopy. Two flat detector technologies prevail. The dominating method is based on an indirect X-ray conversion process, using cesium iodide scintillators. It offers considerable advantages in radiography, angiography and fluoroscopy. The other method employs a direct converter such as selenium which is particularly suitable for mammography. Both flat detector technologies are based on amorphous silicon active pixel matrices. Flat detectors facilitate the clinical workflow in radiographic rooms, foster improved image quality and provide the potential to reduce dose. This added value is based on their large dynamic range, their high sensitivity to X-rays and the instant availability of the image. Advanced image processing is instrumental in these improvements and expand the range of conventional diagnostic methods. In angiography and fluoroscopy the transition from image intensifiers to flat detectors is facilitated by ample advantages they offer, such as distortion-free images, excellent coarse contrast, large dynamic range and high X-ray sensitivity. These characteristics and their compatibility with strong magnetic fields are the basis for improved diagnostic methods and innovative interventional applications. (orig.)

  1. Flat-panel detectors in x-ray diagnosis

    International Nuclear Information System (INIS)

    Spahn, M.; Heer, V.; Freytag, R.

    2003-01-01

    For all application segments X-ray systems with flat-panel detectors increasingly enter the market. In digital radiography, mammography and cardiologic angiography flat-panel detectors are already well established while they are made ready for market introduction in general angiography and fluoroscopy. Two flat-panel detector technologies are available. One technology is based on an indirect conversion process of X-rays while the other one uses a direct conversion method.For radiography and dynamic applications the indirect method provides substantial advantages, while the direct method has some benefits for mammography. In radiography and mammography flat-panel detectors lead to clear improvements with respect to workflow, image quality and dose reduction potentials. These improvements are fostered by the immediate availability of the image, the large dynamic range and the high sensitivity to X-rays. New applications and the use of complex image processing algorithms have the potential to enlarge the present diagnostic range of applications.Up to now, image intensifiers are still the well-established technology for angiography and fluoroscopy. Nevertheless flat-panel detectors begin to enter this field, especially in cardiologic angiography.Characteristics of flat-panel detectors such as the availability of distortion-free images, the excellent contrast resolution, the large dynamic range, the high sensitivity to X-rays and the usability in magnetic fields provide the basis for improved and new diagnostic and interventional methods. (orig.) [de

  2. High precision detector robot arm system

    Science.gov (United States)

    Shu, Deming; Chu, Yong

    2017-01-31

    A method and high precision robot arm system are provided, for example, for X-ray nanodiffraction with an X-ray nanoprobe. The robot arm system includes duo-vertical-stages and a kinematic linkage system. A two-dimensional (2D) vertical plane ultra-precision robot arm supporting an X-ray detector provides positioning and manipulating of the X-ray detector. A vertical support for the 2D vertical plane robot arm includes spaced apart rails respectively engaging a first bearing structure and a second bearing structure carried by the 2D vertical plane robot arm.

  3. Basics principles of flat detector computed tomography (FD-CT)

    International Nuclear Information System (INIS)

    Kyriakou, Y.; Struffert, T.; Doerfler, A.; Kalender, W.A.

    2009-01-01

    Flat detectors (FDs) have been developed for use in radiography and fluoroscopy to replace standard X-ray film, film-screen combinations and image intensifiers (II). In comparison to X-ray film and II, FD technology offers higher dynamic range, dose reduction, fast digital readout and the possibility for dynamic acquisitions of image series, yet keeping to a compact design. It appeared logical to employ FD designs also for computed tomography (CT) imaging. FDCT has meanwhile become widely accepted for interventional and intra-operative imaging using C-arm systems. Additionally, the introduction of FD technology was a milestone for soft-tissue CT imaging in the interventional suite which was not possible with II systems in the past. This review focuses on technical and performance issues of FD technology and its wide range of applications for CT imaging. FDCT is not aimed at challenging standard clinical CT as regards to the typical diagnostic examinations, but it has already proven unique for a number of dedicated CT applications offering distinct practical advantages, above all the availability of immediate CT imaging during an intervention. (orig.) [de

  4. Digital radiography with large-area flat-panel detectors

    International Nuclear Information System (INIS)

    Kotter, E.; Langer, M.

    2002-01-01

    Large-area flat-panel detectors with active readout mechanisms have been on the market for the past 2 years. This article describes different detector technologies. An important distinction is made between detectors with direct and those with indirect conversion of X-rays into electrical charges. Detectors with indirect conversion are built with unstructured or structured scintillators, the latter resulting in less lateral diffusion of emitted light. Some important qualities of flat-panel detectors are discussed. The first phantom and clinical studies published report an image quality at least comparable to that of screen-film systems and a potential for dose reduction. The available studies are summarised in this article. (orig.)

  5. Rocky Flats Neutron Detector Testing at Valduc, France

    International Nuclear Information System (INIS)

    Kim, S.S.; Dulik, G.M.

    2011-01-01

    Recent program requirements of the US Department of Energy/NNSA have led to a need for a criticality accident alarm system to be installed at a newly activated facility. The Criticality Safety Group of the Lawrence Livermore National Laboratory (LLNL) was able to recover and store for possible future use approximately 200 neutron criticality detectors and 20 master alarm panels from the former Rocky Flats Plant in Golden, Colorado when the plant was closed. The Criticality Safety Group participated in a facility analysis and evaluation, the engineering design and review process, as well as the refurbishment, testing, and recalibration of the Rocky Flats criticality alarm system equipment to be used in the new facility. In order to demonstrate the functionality and survivability of the neutron detectors to the effects of an actual criticality accident, neutron detector testing was performed at the French CEA Valduc SILENE reactor from October 7 to October 19, 2010. The neutron detectors were exposed to three criticality events or pulses generated by the SILENE reactor. The first excursion was performed with a bare or unshielded reactor, and the second excursion was made with a lead shielded/reflected reactor, and the third excursion with a polyethylene reflected core. These tests of the Rocky Flats neutron detectors were performed as a part of the 2010 Criticality Accident Alarm System Benchmark Measurements at the SILENE Reactor. The principal investigators for this series of experiments were Thomas M. Miller and John C. Wagner of the Oak Ridge National Laboratory, with Nicolas Authier and Nathalie Baclet of CEA Valduc. Several other organizations were also represented, including the Y-12 National Security Complex, Lawrence Livermore National Laboratory, Los Alamos National Laboratory, CEA Saclay, and Babcock International Group.

  6. The digital flat-panel X-Ray detectors

    International Nuclear Information System (INIS)

    Risticj, S. Goran

    2013-01-01

    In a digital imaging system, the incident x-ray image must be sampled both in the spatial and intensity dimensions. In the spatial dimensions, samples are obtained as averages of the intensity over picture elements or pixels. In the intensity dimension, the signal is digitalized into one of a finite number of levels or bits. Two main types of digital flat-panel detectors are based on the direct conversion, which contains the photoconductor, and on indirect conversion, which contains phosphor. The basics of these detectors are given. Coupling traditional x-ray detection material such as photoconductors and phosphors with a large-area active-matrix readout structure forms the basis of flat panel x-ray images. Active matrix technology provides a new, highly efficient, real time method for electronically storing and measuring the product of the x-ray interaction stage whether the product is visible wavelength photons or electrical charges. The direct and indirect detectors, made as the active-matrix flat-panel detectors containing sensing/storage elements, switching elements (diodes or thin film transistors (TFTS)) and image processing module, are described. Strengths and limitations of stimulable phosphors are discussed. The main advantages and disadvantages of mentioned x-ray detectors are also analyzed. (Author)

  7. Flat-response x-ray-diode-detector development

    International Nuclear Information System (INIS)

    Tirsell, G.

    1982-10-01

    In this report we discuss the design of an improved sub-nanosecond flat response x-ray diode detector needed for ICF diagnostics. This device consists of a high Z cathode and a complex filter tailored to flatten the response so that the total x-ray energy below 1.5 keV can be measured using a single detector. Three major problems have become evident as a result of our work with the original LLNL design including deviation from flatness due to a peak in the response below 200 eV, saturation at relatively low x-ray fluences, and long term gold cathode instability. We are investigating grazing incidence reflection to reduce the response below 200 eV, new high Z cathode materials for long term stability, and a new complex filter for improved flatness. Better saturation performance will require a modified XRD detector under development with reduced anode to cathode spacing and increased anode bias voltage

  8. Present and future of flat panel detectors in the world

    International Nuclear Information System (INIS)

    Inamura, Kiyonari

    2002-01-01

    Present status of development of flat panel detectors and their clinical application in the world have been surveyed, and future trends are also explored especially in the field of material researches and methods of manufacturing. Also the importance of role of medical physicists on user side is described because characteristic physics measurement of a detector assembly is unavoidable and essential in quality assurance in clinical routine and acceptance test in hospitals. Even though physics measurements and clinical evaluations on flat panel detectors have shown remarkable progress and advances in these several years, future problems of cost down in manufacturing and quality assurance to prevent individual differences between detector assemblies must be resolved. Results of evaluation in mammography, chest radiography, fluoroscopy for cardiovascular examination, bone tumor examination and radiotherapy application indicate that flat panel detectors are future promising materials. Their systematic operation is contributing to heighten accuracy of image examinations and preciseness of radiation therapy. Encouragement to medical physicists relevant to flat panel detectors is also raised in this paper. (author)

  9. Study of factors controlling exposure dose and image quality of C-arm in operation room according to detector size of it (Mainly L-Spine AP study)

    Energy Technology Data Exchange (ETDEWEB)

    Chui, Sung Hyun; Jo, Hwang Woo [Dept. of Radiology, Kyung Hee University Hospital at Gangdong, Seoul (Korea, Republic of); Chun, Woon Kwan; Song, Ha Jin [Dept. of Nuclear Engineering, Chosun University, Gwangju (Korea, Republic of); Dong, Kyung Rae [Dept. of Radiological Technology, Gwangju Health University, Gwangju (Korea, Republic of); Choi, Eun Jin [Dept. of Public Health and Medicine, Dongshin University, Naju (Korea, Republic of)

    2015-02-15

    Time of operation has been reduced and accuracy of operation has been improved since C-arm, which offer real-time image of patient, was introduced in operation room. However, because of the contamination of patient, C-arm could not be used more appropriately. Therefore, this study is to know factors of controlling exposure dose, image quality and the exposed dose of health professional in operation room. Height of Wilson frame (bed for operation) was fixed at 130 cm. Then, Model 76-2 Phantom, which was set by assembling manual of Fluke Company, was set on the bed. Head/Spine Fluoroscopy AEC mode was set for exposure condition. According to detector size of C-arm, the absorbed dose per min was measured in the 7 steps OFD (cm) from 10 cm to 40 cm (10, 15, 20, 25, 30, 35, 40 cm). In each step of OFD, the absorbed dose per min of same diameter of collimation was measured. Moreover, using Nero MAX Model 8000, exposure dose per min was measured according to 3 step of distance from detector (20 cm, 60 cm, 100 cm). Finally, resolution was measured by CDRH Disc Phantom and magnification of each OFD was measured by aluminum stick bar. According to detector size of C-arm, difference of absorbed dose shows that the dose of 20 cm OFD is 1.750 times higher than the dose of 40 cm OFD. It means that the C-arm, which has smaller size of detector, shows the bigger difference of absorbed dose per min (p<0.05). In the difference of absorbed dose in the same step of OFD (from 20 cm to 40 cm), the absorbed dose of 9 inch detect or C-arm was 1.370 times higher than 12 inch' s (p<0.05). When OFD was set to 20 cm OFD, the absorbed dose of non-collimation case was approximately 0.816 times lower than the absorbed dose of collimation cases (p<0.05). When the distance was 20 cm from detector, exposed does includes first-ray and scatter-ray. When the distance was 60 cm and 100 cm from detector, exposed does includes just scatter-ray. So, there was the 2.200 times difference of absorbed

  10. Evaluation of a flat-panel detector system

    International Nuclear Information System (INIS)

    Sato, Masami; Eguchi, Yoichi; Yamada, Kinichi; Kaga, Yuji; Endo, Yutaka; Yamazaki, Tatsuya

    2001-01-01

    We evaluated the imaging performance of a flat-panel detector digital radiography system (CXDI-11 X-ray Digital Camera, Canon Inc.) and a computed radiography system (FCR9000C-HQ, Fuji Film). The characteristics of the two detectors and of the overall systems were compared. This included evaluation and comparison of the fundamental physical characteristics, including x-ray response curve, modulation transfer function (MTF), Wiener spectra, noise-equivalent quanta, and x-ray tube voltage-dependent detector response. Overall system performance was evaluated using receiver operating characteristic (ROC) analysis. The results of the study showed that the dynamic range of the CXDI-11 measured relative to the input x-ray flux was 10 3 , similar to that of the FCR9000C-HQ. Both systems showed similar final MTFs, although the pre-sampling MTF of the CXDI-11 was better than that of the FCR9000C-HQ. Noise analysis, based on noise-equivalent quanta and Wiener spectra, showed that for normal exposure conditions the CXDI-11 had superior performance. With both systems, x-ray response (system output/incident x-ray exposure) increased with increasing x-ray tube voltage. ROC analysis indicated that the CXDI-11 was superior in overall performance. (author)

  11. Performance evaluation of flat panel detector in x-ray fluoroscopy

    International Nuclear Information System (INIS)

    Grewal, R.K.; Mclean, I.D.

    2004-01-01

    Full text: Flat panel detectors are currently replacing the conventional image intensifiers in R-F imaging. We evaluated the performance of a biplane cardiac imaging system (Siemens Axiom Artis dBC), the image acquisition was based on a 25 cm diagonal digital fiat panel detector. Performance characteristics included image quality, typical patient entrance dose and measurement of input to the surface of flat detector. The results were compared with conventional image intensifier systems (Siemens Hicor Unit and Toshiba DPF 2000 A Biplane Unit) used in cardiac imaging at Westmead. Image quality and dose measurements were performed following standard protocols using Westmead test object and 20 cm solid water as absorber in the beam. For measurement of input to the surface of flat detector, 2 mm copper was placed on the collimator. Radcal 3cc and 180 cc ion chambers were used for dose measurements. Image quality: Our measurements on flat panel system indicate that high contrast resolution and threshold contrast is not affected by changing field size. This is expected due to minimum loss of signal in the imaging chain of digital systems and the independence of detector pixel size with change in field of view. While low contrast resolution was found to be similar to conventional systems, high contrast resolution was significantly superior using flat detector system for large and intermediate field of view (25-28 1p/cm against 18-20). Typical patient dose as measured using flat detector system was similar to the conventional Toshiba pulsed fluoroscopy system( ∼ 3 - 8 mGy/min depending on the field size). This was 40-50 % lower than our old Siemens hicore unit. Input to the surface of flat detector was found to vary with field size as is the case with a conventional II system. As described elsewhere, although there is no necessity to increase exposure or video gain in a digital magnification, digital data interpolation process introduces noise. As a result system

  12. Evaluation of arm-leg coordination in flat breaststroke.

    Science.gov (United States)

    Chollet, D; Seifert, L; Leblanc, H; Boulesteix, L; Carter, M

    2004-10-01

    This study proposes a new method to evaluate arm-leg coordination in flat breaststroke. Five arm and leg stroke phases were defined with a velocity-video system. Five time gaps quantified the time between arm and leg actions during three paces of a race (200 m, 100 m and 50 m) in 16 top level swimmers. Based on these time gaps, effective glide, effective propulsion, effective leg insweep and effective recovery were used to identify the different stroke phases of the body. A faster pace corresponded to increased stroke rate, decreased stroke length, increased propulsive phases, shorter glide phases, and a shorter T1 time gap, which measured the effective body glide. The top level swimmers showed short time gaps (T2, T3, T4, measuring the timing of arm-leg recoveries), which reflected the continuity in arm and leg actions. The measurement of these time gaps thus provides a pertinent evaluation of swimmers' skill in adapting their arm-leg coordination to biomechanical constraints.

  13. Flat-screen detector systems in skeletal radiology

    International Nuclear Information System (INIS)

    Grampp, S.; Czerny, C.; Krestan, C.; Henk, C.; Heiner, L.; Imhof, H.

    2003-01-01

    Implementation of flat-panel detectors and digital integration of the technique instead of the use of conventional radiographs leads to a shortening of the work process. With flat-panel technology the image production process is shortened by more than 30%. Major advantages in the implementation of integrated RIS, PACS and flat-panel detector system are increases in quality because most mistakes in picture labeling can be avoided, easier handling without the need for cassettes, and the possibility of image post-processing. The diagnostic quality of the images in the field of musculoskeletal radiology is, in comparison to conventional radiographs, at least adequate and in most cases markedly improved with a marked reduction in radiation exposure of around 30-50%. With respect to the numerous advantages of the digital techniques and especially flat-panel technology there is a very high likelihood that conventional radiographs will be substituted in the coming years, even though the cost of the new technology is currently significantly higher compared to conventional systems. (orig.) [de

  14. Concepts for dose determination in flat-detector CT

    Science.gov (United States)

    Kyriakou, Yiannis; Deak, Paul; Langner, Oliver; Kalender, Willi A.

    2008-07-01

    Flat-detector computed tomography (FD-CT) scanners provide large irradiation fields of typically 200 mm in the cranio-caudal direction. In consequence, dose assessment according to the current definition of the computed tomography dose index CTDIL=100 mm, where L is the integration length, would demand larger ionization chambers and phantoms which do not appear practical. We investigated the usefulness of the CTDI concept and practical dosimetry approaches for FD-CT by measurements and Monte Carlo (MC) simulations. An MC simulation tool (ImpactMC, VAMP GmbH, Erlangen, Germany) was used to assess the dose characteristics and was calibrated with measurements of air kerma. For validation purposes measurements were performed on an Axiom Artis C-arm system (Siemens Medical Solutions, Forchheim, Germany) equipped with a flat detector of 40 cm × 30 cm. The dose was assessed for 70 kV and 125 kV in cylindrical PMMA phantoms of 160 mm and 320 mm diameter with a varying phantom length from 150 to 900 mm. MC simulation results were compared to the values obtained with a calibrated ionization chambers of 100 mm and 250 mm length and to thermoluminesence (TLD) dose profiles. The MCs simulations were used to calculate the efficiency of the CTDIL determination with respect to the desired CTDI∞. Both the MC simulation results and the dose distributions obtained by MC simulation were in very good agreement with the CTDI measurements and with the reference TLD profiles, respectively, to within 5%. Standard CTDI phantoms which have a z-extent of 150 mm underestimate the dose at the center by up to 55%, whereas a z-extent of >=600 mm appears to be sufficient for FD-CT; the baseline value of the respective profile was within 1% to the reference baseline. As expected, the measurements with ionization chambers of 100 mm and 250 mm offer a limited accuracy, whereas an increased integration length of >=600 mm appeared to be necessary to approximate CTDI∞ in within 1%. MC simulations

  15. Ray tracing reconstruction investigation for C-arm tomosynthesis

    Science.gov (United States)

    Malalla, Nuhad A. Y.; Chen, Ying

    2016-04-01

    C-arm tomosynthesis is a three dimensional imaging technique. Both x-ray source and the detector are mounted on a C-arm wheeled structure to provide wide variety of movement around the object. In this paper, C-arm tomosynthesis was introduced to provide three dimensional information over a limited view angle (less than 180o) to reduce radiation exposure and examination time. Reconstruction algorithms based on ray tracing method such as ray tracing back projection (BP), simultaneous algebraic reconstruction technique (SART) and maximum likelihood expectation maximization (MLEM) were developed for C-arm tomosynthesis. C-arm tomosynthesis projection images of simulated spherical object were simulated with a virtual geometric configuration with a total view angle of 40 degrees. This study demonstrated the sharpness of in-plane reconstructed structure and effectiveness of removing out-of-plane blur for each reconstruction algorithms. Results showed the ability of ray tracing based reconstruction algorithms to provide three dimensional information with limited angle C-arm tomosynthesis.

  16. C-arm flat-panel CT arthrography of the shoulder: Radiation dose considerations and preliminary data on diagnostic performance

    International Nuclear Information System (INIS)

    Guggenberger, Roman; Ulbrich, Erika J.; Kaelin, Pascal; Pfammatter, Thomas; Alkadhi, Hatem; Andreisek, Gustav; Dietrich, Tobias J.; Scholz, Rosemarie; Koehler, Christoph; Elsaesser, Thilo; Le Corroller, Thomas

    2017-01-01

    To investigate radiation dose and diagnostic performance of C-arm flat-panel CT (FPCT) versus standard multi-detector CT (MDCT) shoulder arthrography using MRI-arthrography as reference standard. Radiation dose of two different FPCT acquisitions (5 and 20 s) and standard MDCT of the shoulder were assessed using phantoms and thermoluminescence dosimetry. FPCT arthrographies were performed in 34 patients (mean age 44 ± 15 years). Different joint structures were quantitatively and qualitatively assessed by two independent radiologists. Inter-reader agreement and diagnostic performance were calculated. Effective radiation dose was markedly lower in FPCT 5 s (0.6 mSv) compared to MDCT (1.7 mSv) and FPCT 20 s (3.4 mSv). Contrast-to-noise ratios (CNRs) were significantly (p < 0.05) higher in FPCT 20-s versus 5-s protocols. Inter-reader agreements of qualitative ratings ranged between κ = 0.47-1.0. Sensitivities for cartilage and rotator cuff pathologies were low for FPCT 5-s (40 % and 20 %) and moderate for FPCT 20-s protocols (75 % and 73 %). FPCT showed high sensitivity (81-86 % and 89-99 %) for bone and acromioclavicular-joint pathologies. Using a 5-s protocol FPCT shoulder arthrography provides lower radiation dose compared to MDCT but poor sensitivity for cartilage and rotator cuff pathologies. FPCT 20-s protocol is moderately sensitive for cartilage and rotator cuff tendon pathology with markedly higher radiation dose compared to MDCT. (orig.)

  17. C-arm flat-panel CT arthrography of the shoulder: Radiation dose considerations and preliminary data on diagnostic performance

    Energy Technology Data Exchange (ETDEWEB)

    Guggenberger, Roman; Ulbrich, Erika J.; Kaelin, Pascal; Pfammatter, Thomas; Alkadhi, Hatem; Andreisek, Gustav [University Hospital Zurich, Institute of Diagnostic and Interventional Radiology, Zuerich (Switzerland); Dietrich, Tobias J. [Balgrist University Hospital, Department of Radiology, Zurich (Switzerland); Scholz, Rosemarie; Koehler, Christoph; Elsaesser, Thilo [Siemens Healthcare GmbH, Business Area Advanced Therapies, Forchheim (Germany); Le Corroller, Thomas [Aix-Marseille Universite, CNRS, ISM UMR 7287, Marseille (France); Radiology Department, APHM, Marseille (France)

    2017-02-15

    To investigate radiation dose and diagnostic performance of C-arm flat-panel CT (FPCT) versus standard multi-detector CT (MDCT) shoulder arthrography using MRI-arthrography as reference standard. Radiation dose of two different FPCT acquisitions (5 and 20 s) and standard MDCT of the shoulder were assessed using phantoms and thermoluminescence dosimetry. FPCT arthrographies were performed in 34 patients (mean age 44 ± 15 years). Different joint structures were quantitatively and qualitatively assessed by two independent radiologists. Inter-reader agreement and diagnostic performance were calculated. Effective radiation dose was markedly lower in FPCT 5 s (0.6 mSv) compared to MDCT (1.7 mSv) and FPCT 20 s (3.4 mSv). Contrast-to-noise ratios (CNRs) were significantly (p < 0.05) higher in FPCT 20-s versus 5-s protocols. Inter-reader agreements of qualitative ratings ranged between κ = 0.47-1.0. Sensitivities for cartilage and rotator cuff pathologies were low for FPCT 5-s (40 % and 20 %) and moderate for FPCT 20-s protocols (75 % and 73 %). FPCT showed high sensitivity (81-86 % and 89-99 %) for bone and acromioclavicular-joint pathologies. Using a 5-s protocol FPCT shoulder arthrography provides lower radiation dose compared to MDCT but poor sensitivity for cartilage and rotator cuff pathologies. FPCT 20-s protocol is moderately sensitive for cartilage and rotator cuff tendon pathology with markedly higher radiation dose compared to MDCT. (orig.)

  18. Radiation exposure to operating staff during rotational flat-panel angiography and C-arm cone beam computed tomography (CT) applications

    Energy Technology Data Exchange (ETDEWEB)

    Schulz, Boris, E-mail: boris.schell@googlemail.com [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Heidenreich, Ralf, E-mail: ralf.heidenreich@roentgen-consult.de [Röntgen-Consult Company, Schulhausstrasse 37, 79199 Kirchzarten (Germany); Heidenreich, Monika, E-mail: info@roentgen-consult.de [Röntgen-Consult Company, Schulhausstrasse 37, 79199 Kirchzarten (Germany); Eichler, Katrin, E-mail: k.eichler@em.uni-frankfurt.de [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Thalhammer, Axel, E-mail: axel.thalhammer@kgu.de [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Naeem, Naguib Nagy Naguib, E-mail: nagynnn@yahoo.com [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Vogl, Thomas Josef, E-mail: T.Vogl@em.uni-frankfurt.de [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Zangos, Stefan, E-mail: Zangos@em.uni-frankfurt.de [Goethe University Hospital, Department of Diagnostic and Interventional Radiology, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany)

    2012-12-15

    Purpose: To evaluate the radiation exposure for operating personel associated with rotational flat-panel angiography and C-arm cone beam CT. Materials and methods: Using a dedicated angiography-suite, 2D and 3D examinations of the liver were performed on a phantom to generate scattered radiation. Exposure was measured with a dosimeter at predefined heights (eye, thyroid, breast, gonads and knee) at the physician's location. Analysis included 3D procedures with a field of view (FOV) of 24 cm × 18 cm (8 s/rotation, 20 s/rotation and 5 s/2 rotations), and 47 cm × 18 cm (16 s/2 rotations) and standard 2D angiography (10 s, FOV 24 cm × 18 cm). Results: Measurements showed the highest radiation dose at the eye and thyroid level. In comparison to 2D-DSA (3.9 μSv eye-exposure), the 3D procedures caused an increased radiation exposure both in standard FOV (8 s/rotation: 28.0 μSv, 20 s/rotation: 79.3 μSv, 5 s/2 rotations: 32.5 μSv) and large FOV (37.6 μSv). Proportional distributions were measured for the residual heights. With the use of lead glass, irradiation of the eye lens was reduced to 0.2 μSv (2D DSA) and 10.6 μSv (3D technique with 20 s/rotation). Conclusion: Rotational flat-panel angiography and C-arm cone beam applications significantly increase radiation exposure to the attending operator in comparison to 2D angiography. Our study indicates that the physician should wear protective devices and leave the examination room when performing 3D examinations.

  19. Radiation exposure to operating staff during rotational flat-panel angiography and C-arm cone beam computed tomography (CT) applications.

    Science.gov (United States)

    Schulz, Boris; Heidenreich, Ralf; Heidenreich, Monika; Eichler, Katrin; Thalhammer, Axel; Naeem, Naguib Nagy Naguib; Vogl, Thomas Josef; Zangos, Stefan

    2012-12-01

    To evaluate the radiation exposure for operating personnel associated with rotational flat-panel angiography and C-arm cone beam CT. Using a dedicated angiography-suite, 2D and 3D examinations of the liver were performed on a phantom to generate scattered radiation. Exposure was measured with a dosimeter at predefined heights (eye, thyroid, breast, gonads and knee) at the physician's location. Analysis included 3D procedures with a field of view (FOV) of 24 cm × 18 cm (8s/rotation, 20s/rotation and 5s/2 rotations), and 47 cm×18 cm (16s/2 rotations) and standard 2D angiography (10s, FOV 24 cm×18 cm). Measurements showed the highest radiation dose at the eye and thyroid level. In comparison to 2D-DSA (3.9 μSv eye-exposure), the 3D procedures caused an increased radiation exposure both in standard FOV (8s/rotation: 28.0 μSv, 20s/rotation: 79.3 μSv, 5s/2 rotations: 32.5 μSv) and large FOV (37.6 μSv). Proportional distributions were measured for the residual heights. With the use of lead glass, irradiation of the eye lens was reduced to 0.2 μSv (2D DSA) and 10.6 μSv (3D technique with 20s/rotation). Rotational flat-panel angiography and C-arm cone beam applications significantly increase radiation exposure to the attending operator in comparison to 2D angiography. Our study indicates that the physician should wear protective devices and leave the examination room when performing 3D examinations. Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

  20. Radiation exposure to operating staff during rotational flat-panel angiography and C-arm cone beam computed tomography (CT) applications

    International Nuclear Information System (INIS)

    Schulz, Boris; Heidenreich, Ralf; Heidenreich, Monika; Eichler, Katrin; Thalhammer, Axel; Naeem, Naguib Nagy Naguib; Vogl, Thomas Josef; Zangos, Stefan

    2012-01-01

    Purpose: To evaluate the radiation exposure for operating personel associated with rotational flat-panel angiography and C-arm cone beam CT. Materials and methods: Using a dedicated angiography-suite, 2D and 3D examinations of the liver were performed on a phantom to generate scattered radiation. Exposure was measured with a dosimeter at predefined heights (eye, thyroid, breast, gonads and knee) at the physician's location. Analysis included 3D procedures with a field of view (FOV) of 24 cm × 18 cm (8 s/rotation, 20 s/rotation and 5 s/2 rotations), and 47 cm × 18 cm (16 s/2 rotations) and standard 2D angiography (10 s, FOV 24 cm × 18 cm). Results: Measurements showed the highest radiation dose at the eye and thyroid level. In comparison to 2D-DSA (3.9 μSv eye-exposure), the 3D procedures caused an increased radiation exposure both in standard FOV (8 s/rotation: 28.0 μSv, 20 s/rotation: 79.3 μSv, 5 s/2 rotations: 32.5 μSv) and large FOV (37.6 μSv). Proportional distributions were measured for the residual heights. With the use of lead glass, irradiation of the eye lens was reduced to 0.2 μSv (2D DSA) and 10.6 μSv (3D technique with 20 s/rotation). Conclusion: Rotational flat-panel angiography and C-arm cone beam applications significantly increase radiation exposure to the attending operator in comparison to 2D angiography. Our study indicates that the physician should wear protective devices and leave the examination room when performing 3D examinations.

  1. C-arm computed tomography for transarterial chemoperfusion and chemo-embolization of thoracic lesions; Transarterielle Chemoperfusion und -embolisation thorakaler Neoplasmen mittels C-Arm CT

    Energy Technology Data Exchange (ETDEWEB)

    Vogl, T.J.; Naguib, N.N.; Nour-Eldin, N.E.; Lehnert, T.; Mbalisike, E. [Klinikum der Johann-Wolfgang-Goethe-Universitaet, Institut fuer Diagnostische und Interventionelle Radiologie, Frankfurt am Main (Germany)

    2009-09-15

    To evaluate the role of C-arm CT for on-line fluoroscopy in regional transarterial chemoperfusion (TACP) and chemo-embolization (TPCE) of primary and secondary malignant thoracic lesions. From September 2008 to March 2009 a total of 31 patients (20 males and 11 females, average age: 61.7 years, range 22-84 years) with 53 thoracic malignant lesions from different origins (primary or secondary pulmonary carcinoma n=37, pleural mesothelioma n=16) were treated with TACP or TPCE using flat-detector CT (FD-CT). C-arm CT of the latest generation was used to localize the lesion before local chemotherapy (Artis Zeego, Siemens, Erlangen). For TACP a 220 rotation and a volume of 150 ml (ratio of 1:2 contrast/normal saline), delay 2 s and flow 12 ml/s was used. For TPCE a volume of 75 ml (ratio of 1:2 contrast/normal saline), delay 2 s and flow 3 ml/s was used. TPCE C-arm CT allowed the evaluation of the degree of perfusion of the tumor and the geographic areas of enhancement correlated with the post-interventional lipiodol uptake in MSCT. In TACP the intercostal arteries involved could be visualized and in 30% of interventions the catheter had to be repositioned for the following intervention. C-arm CT provides additional information on the vascular characteristics and perfusion of pulmonary lesions resulting in a change of interventional strategy in a relevant number of patients. (orig.) [German] Ziel der Arbeit war die Evaluation der Wertigkeit der C-Arm CT fuer die online gesteuerte regionale transarterielle Chemoperfusion (TACP) und die transpulmonale Chemoembolisation (TPCE) primaerer und sekundaerer thorakaler Neoplasmen Von September 2008 bis Maerz 2009 wurden 31 Patienten (11 Frauen/20 Maenner, Durchschnittsalter 61,7 Jahre) mit 53 unterschiedlichen thorakalen Neoplasmen (primaere oder sekundaere Lungenkarzinome [n=37], Pleuramesotheliome [n=16]) mittels TACP oder TPCE unter Einsatz der Flachdetektortechnologie (FD-CT) behandelt. Alle Behandlungen erfolgten an einem C

  2. Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery

    International Nuclear Information System (INIS)

    Petersen, Asger Greval; Eiskjaer, Soeren; Kaspersen, Jon

    2012-01-01

    During surgery for spinal deformities, accurate placement of pedicle screws may be guided by intraoperative cone-beam flat-detector CT. The purpose of this study was to identify appropriate paediatric imaging protocols aiming to reduce the radiation dose in line with the ALARA principle. Using O-arm registered (Medtronic, Inc.), three paediatric phantoms were employed to measure CTDI w doses with default and lowered exposure settings. Images from 126 scans were evaluated by two spinal surgeons and scores were compared (Kappa statistics). Effective doses were calculated. The recommended new low-dose 3-D spine protocols were then used in 15 children. The lowest acceptable exposure as judged by image quality for intraoperative use was 70 kVp/40 mAs, 70 kVp/80 mAs and 80 kVp/40 mAs for the 1-, 5- and 12-year-old-equivalent phantoms respectively (kappa = 0,70). Optimised dose settings reduced CTDI w doses 89-93%. The effective dose was 0.5 mSv (91-94,5% reduction). The optimised protocols were used clinically without problems. Radiation doses for intraoperative 3-D CT using a cone-beam flat-detector scanner could be reduced at least 89% compared to manufacturer settings and still be used to safely navigate pedicle screws. (orig.)

  3. Concepts for dose determination in flat-detector CT

    International Nuclear Information System (INIS)

    Kyriakou, Yiannis; Deak, Paul; Langner, Oliver; Kalender, Willi A

    2008-01-01

    Flat-detector computed tomography (FD-CT) scanners provide large irradiation fields of typically 200 mm in the cranio-caudal direction. In consequence, dose assessment according to the current definition of the computed tomography dose index CTDI L=100mm , where L is the integration length, would demand larger ionization chambers and phantoms which do not appear practical. We investigated the usefulness of the CTDI concept and practical dosimetry approaches for FD-CT by measurements and Monte Carlo (MC) simulations. An MC simulation tool (ImpactMC, VAMP GmbH, Erlangen, Germany) was used to assess the dose characteristics and was calibrated with measurements of air kerma. For validation purposes measurements were performed on an Axiom Artis C-arm system (Siemens Medical Solutions, Forchheim, Germany) equipped with a flat detector of 40 cm x 30 cm. The dose was assessed for 70 kV and 125 kV in cylindrical PMMA phantoms of 160 mm and 320 mm diameter with a varying phantom length from 150 to 900 mm. MC simulation results were compared to the values obtained with a calibrated ionization chambers of 100 mm and 250 mm length and to thermoluminesence (TLD) dose profiles. The MCs simulations were used to calculate the efficiency of the CTDI L determination with respect to the desired CTDI ∞ . Both the MC simulation results and the dose distributions obtained by MC simulation were in very good agreement with the CTDI measurements and with the reference TLD profiles, respectively, to within 5%. Standard CTDI phantoms which have a z-extent of 150 mm underestimate the dose at the center by up to 55%, whereas a z-extent of ≥600 mm appears to be sufficient for FD-CT; the baseline value of the respective profile was within 1% to the reference baseline. As expected, the measurements with ionization chambers of 100 mm and 250 mm offer a limited accuracy, whereas an increased integration length of ≥600 mm appeared to be necessary to approximate CTDI ∞ in within 1%. MC

  4. Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws.

    Science.gov (United States)

    Filli, Lukas; Marcon, Magda; Scholz, Bernhard; Calcagni, Maurizio; Finkenstädt, Tim; Andreisek, Gustav; Guggenberger, Roman

    2014-12-01

    The aim of this study was to evaluate a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography (FDCT) of scaphoid fixation screws. FDCT has gained interest in imaging small anatomic structures of the appendicular skeleton. Angiographic C-arm systems with flat detectors allow fluoroscopy and FDCT imaging in a one-stop procedure emphasizing their role as an ideal intraoperative imaging tool. However, FDCT imaging can be significantly impaired by artefacts induced by fixation screws. Following ethical board approval, commercially available scaphoid fixation screws were inserted into six cadaveric specimens in order to fix artificially induced scaphoid fractures. FDCT images corrected with the algorithm were compared to uncorrected images both quantitatively and qualitatively by two independent radiologists in terms of artefacts, screw contour, fracture line visibility, bone visibility, and soft tissue definition. Normal distribution of variables was evaluated using the Kolmogorov-Smirnov test. In case of normal distribution, quantitative variables were compared using paired Student's t tests. The Wilcoxon signed-rank test was used for quantitative variables without normal distribution and all qualitative variables. A p value of < 0.05 was considered to indicate statistically significant differences. Metal artefacts were significantly reduced by the correction algorithm (p < 0.001), and the fracture line was more clearly defined (p < 0.01). The inter-observer reliability was "almost perfect" (intra-class correlation coefficient 0.85, p < 0.001). The prototype correction algorithm in FDCT for metal artefacts induced by scaphoid fixation screws may facilitate intra- and postoperative follow-up imaging. Flat detector computed tomography (FDCT) is a helpful imaging tool for scaphoid fixation. The correction algorithm significantly reduces artefacts in FDCT induced by scaphoid fixation screws. This may facilitate intra

  5. Reconstruction of Spectra Using X-ray Flat Panel Detector; Reconstruccion de Espectros de Rayos X Utilizando un Detector Flat Panel

    Energy Technology Data Exchange (ETDEWEB)

    Gallardo, S.; Querol, A.; Pozuelo, F.; Juste, B.; Rodenas, J.; Verdu, G.

    2013-07-01

    In this work, we used a flat panel detector with a wedge of PMMA for absorbed dose curve for given working conditions of X-ray tube The relationship between absorbed dose curve recorded by the flat panel and primary X-ray spectrum is defined by a response function that can be obtained using the Monte Carlo method, namely the MCNP5 code. However there are some problems that affect the applicability of this method such as: flat panel characteristics and the characteristics of the physical process (ill-conditioned problem). Both aspects are discussed in this paper.

  6. Ultra-high-resolution C-arm flat-detector CT angiography evaluation reveals 3-fold higher association rate for sporadic intracranial cavernous malformations and developmental venous anomalies: a retrospective study in consecutive 58 patients with 60 cavernous malformations

    Energy Technology Data Exchange (ETDEWEB)

    Kocak, Burak [Aksaray State Hospital, Department of Radiology, Aksaray (Turkey); Kizilkilic, Osman; Kocer, Naci; Islak, Civan [Istanbul University, Department of Radiology, Cerrahpasa Medical Faculty, Istanbul (Turkey); Oz, Buge; Bakkaloglu, Dogu Vuralli [Istanbul University, Department of Pathology, Cerrahpasa Medical Faculty, Istanbul (Turkey); Isler, Cihan [Istanbul University, Department of Neurosurgery, Cerrahpasa Medical Faculty, Istanbul (Turkey)

    2017-06-15

    The imaging and surgical literature has confusing association rates for the association between sporadic intracranial cavernous malformations (CMs) and developmental venous anomalies (DVAs). In this study, our purpose was to determine the association rate using ultra-high-resolution C-arm flat-detector CT angiography (FDCTA) and compare it with literature. Fifty-eight patients with 60 sporadic intracranial CMs that underwent an FDCTA study were included in our retrospective study. Re-evaluation of radiological data was performed based on the criteria defined by authors. Isotropic volumetric reconstructions with ultra-high resolution (voxel size of 102 μm{sup 3} for initial; 67 μm{sup 3} and 32 μm{sup 3} for further evaluation) were used for assessment. Sixteen patients underwent surgery for excision of their CMs. Fifty-one of all patients (87.9 %) were associated with a DVA. Undefined local venous structures (UD-LVSs) were observed in the remaining 7 patients (12.1 %). The strength of interobserver agreement was excellent [kappa(k) coefficient = 0.923]. Ultra-high-resolution FDCTA evaluation of CMs and DVAs reveals 3-fold higher association rate compared to the literature. FDCTA for patients with sporadic CMs could help identify the associated DVAs that remained undetected or unclear with other imaging modalities, which can be useful in decision-making processes, planning surgery, and during operation. (orig.)

  7. Modelling of scintillator based flat-panel detectors with Monte-Carlo simulations

    International Nuclear Information System (INIS)

    Reims, N; Sukowski, F; Uhlmann, N

    2011-01-01

    Scintillator based flat panel detectors are state of the art in the field of industrial X-ray imaging applications. Choosing the proper system and setup parameters for the vast range of different applications can be a time consuming task, especially when developing new detector systems. Since the system behaviour cannot always be foreseen easily, Monte-Carlo (MC) simulations are keys to gain further knowledge of system components and their behaviour for different imaging conditions. In this work we used two Monte-Carlo based models to examine an indirect converting flat panel detector, specifically the Hamamatsu C9312SK. We focused on the signal generation in the scintillation layer and its influence on the spatial resolution of the whole system. The models differ significantly in their level of complexity. The first model gives a global description of the detector based on different parameters characterizing the spatial resolution. With relatively small effort a simulation model can be developed which equates the real detector regarding signal transfer. The second model allows a more detailed insight of the system. It is based on the well established cascade theory, i.e. describing the detector as a cascade of elemental gain and scattering stages, which represent the built in components and their signal transfer behaviour. In comparison to the first model the influence of single components especially the important light spread behaviour in the scintillator can be analysed in a more differentiated way. Although the implementation of the second model is more time consuming both models have in common that a relatively small amount of system manufacturer parameters are needed. The results of both models were in good agreement with the measured parameters of the real system.

  8. Accuracy of x-ray image-based 3D localization from two C-arm views: a comparison between an ideal system and a real device

    Science.gov (United States)

    Brost, Alexander; Strobel, Norbert; Yatziv, Liron; Gilson, Wesley; Meyer, Bernhard; Hornegger, Joachim; Lewin, Jonathan; Wacker, Frank

    2009-02-01

    arm X-ray imaging devices are commonly used for minimally invasive cardiovascular or other interventional procedures. Calibrated state-of-the-art systems can, however, not only be used for 2D imaging but also for three-dimensional reconstruction either using tomographic techniques or even stereotactic approaches. To evaluate the accuracy of X-ray object localization from two views, a simulation study assuming an ideal imaging geometry was carried out first. This was backed up with a phantom experiment involving a real C-arm angiography system. Both studies were based on a phantom comprising five point objects. These point objects were projected onto a flat-panel detector under different C-arm view positions. The resulting 2D positions were perturbed by adding Gaussian noise to simulate 2D point localization errors. In the next step, 3D point positions were triangulated from two views. A 3D error was computed by taking differences between the reconstructed 3D positions using the perturbed 2D positions and the initial 3D positions of the five points. This experiment was repeated for various C-arm angulations involving angular differences ranging from 15° to 165°. The smallest 3D reconstruction error was achieved, as expected, by views that were 90° degrees apart. In this case, the simulation study yielded a 3D error of 0.82 mm +/- 0.24 mm (mean +/- standard deviation) for 2D noise with a standard deviation of 1.232 mm (4 detector pixels). The experimental result for this view configuration obtained on an AXIOM Artis C-arm (Siemens AG, Healthcare Sector, Forchheim, Germany) system was 0.98 mm +/- 0.29 mm, respectively. These results show that state-of-the-art C-arm systems can localize instruments with millimeter accuracy, and that they can accomplish this almost as well as an idealized theoretical counterpart. High stereotactic localization accuracy, good patient access, and CT-like 3D imaging capabilities render state-of-the-art C-arm systems ideal devices for X

  9. A semiempirical linear model of indirect, flat-panel x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M. [Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Psychological and Brain Sciences, University of California, Santa Barbara, California 92106 (United States); Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 3100, Sacramento, California 95817 (United States)

    2012-04-15

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r

  10. A semiempirical linear model of indirect, flat-panel x-ray detectors

    International Nuclear Information System (INIS)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M.

    2012-01-01

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r 2 of

  11. A semiempirical linear model of indirect, flat-panel x-ray detectors.

    Science.gov (United States)

    Huang, Shih-Ying; Yang, Kai; Abbey, Craig K; Boone, John M

    2012-04-01

    It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV∕0.1-mm Sn, 65 kV∕0.2-mm Cu, 85 kV∕1.5-mm Al, and 95 kV∕0.05-mm Ag. The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r(2) of signal mean to k

  12. Proton tagging with the one arm AFP detector

    CERN Document Server

    The ATLAS collaboration

    2017-01-01

    The ATLAS Forward Proton (AFP) detector is designed to identify events in which one or two protons emerge intact from the LHC collisions at small angles. Such processes are usually associated with diffractive scattering. The first arm of the detector, consisting of tracking planes of 3D silicon sensors, was installed in February 2016. This note presents detector level distributions of events with a tagged proton with the AFP tracker and jets in the central detector in the final state.

  13. Synchrotron applications of an amorphous silicon flat-panel detector

    International Nuclear Information System (INIS)

    Lee, J. H.; Can Aydiner, C.; Almer, J.; Bernier, J.; Chapman, K. W.; Chupas, P. J.; Haeffner, D.; Kump, K.; Lee, P. L.; Lienert, U.; Miceli, A.; Vera, G.; LANL; GE Healthcare

    2008-01-01

    A GE Revolution 41RT flat-panel detector (GE 41RT) from GE Healthcare (GE) has been in operation at the Advanced Photon Source for over two years. The detector has an active area of 41 cm x 41 cm with 200 (micro)m x 200 (micro)m pixel size. The nominal working photon energy is around 80 keV. The physical set-up and utility software of the detector system are discussed in this article. The linearity of the detector response was measured at 80.7 keV. The memory effect of the detector element, called lag, was also measured at different exposure times and gain settings. The modulation transfer function was measured in terms of the line-spread function using a 25 (micro)m x 1 cm tungsten slit. The background (dark) signal, the signal that the detector will carry without exposure to X-rays, was measured at three different gain settings and with exposure times of 1 ms to 15 s. The radial geometric flatness of the sensor panel was measured using the diffraction pattern from a CeO 2 powder standard. The large active area and fast data-capturing rate, i.e. 8 frames s -1 in radiography mode, 30 frames s -1 in fluoroscopy mode, make the GE 41RT one of a kind and very versatile in synchrotron diffraction. The loading behavior of a Cu/Nb multilayer material is used to demonstrate the use of the detector in a strain-stress experiment. Data from the measurement of various samples, amorphous SiO 2 in particular, are presented to show the detector effectiveness in pair distribution function measurements

  14. Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws

    Energy Technology Data Exchange (ETDEWEB)

    Filli, Lukas; Finkenstaedt, Tim; Andreisek, Gustav; Guggenberger, Roman [University Hospital of Zurich, Department of Diagnostic and Interventional Radiology, Zurich (Switzerland); Marcon, Magda [University Hospital of Zurich, Department of Diagnostic and Interventional Radiology, Zurich (Switzerland); University of Udine, Institute of Diagnostic Radiology, Department of Medical and Biological Sciences, Udine (Italy); Scholz, Bernhard [Imaging and Therapy Division, Siemens AG, Healthcare Sector, Forchheim (Germany); Calcagni, Maurizio [University Hospital of Zurich, Division of Plastic Surgery and Hand Surgery, Zurich (Switzerland)

    2014-12-15

    The aim of this study was to evaluate a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography (FDCT) of scaphoid fixation screws. FDCT has gained interest in imaging small anatomic structures of the appendicular skeleton. Angiographic C-arm systems with flat detectors allow fluoroscopy and FDCT imaging in a one-stop procedure emphasizing their role as an ideal intraoperative imaging tool. However, FDCT imaging can be significantly impaired by artefacts induced by fixation screws. Following ethical board approval, commercially available scaphoid fixation screws were inserted into six cadaveric specimens in order to fix artificially induced scaphoid fractures. FDCT images corrected with the algorithm were compared to uncorrected images both quantitatively and qualitatively by two independent radiologists in terms of artefacts, screw contour, fracture line visibility, bone visibility, and soft tissue definition. Normal distribution of variables was evaluated using the Kolmogorov-Smirnov test. In case of normal distribution, quantitative variables were compared using paired Student's t tests. The Wilcoxon signed-rank test was used for quantitative variables without normal distribution and all qualitative variables. A p value of < 0.05 was considered to indicate statistically significant differences. Metal artefacts were significantly reduced by the correction algorithm (p < 0.001), and the fracture line was more clearly defined (p < 0.01). The inter-observer reliability was ''almost perfect'' (intra-class correlation coefficient 0.85, p < 0.001). The prototype correction algorithm in FDCT for metal artefacts induced by scaphoid fixation screws may facilitate intra- and postoperative follow-up imaging. (orig.)

  15. Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws

    International Nuclear Information System (INIS)

    Filli, Lukas; Finkenstaedt, Tim; Andreisek, Gustav; Guggenberger, Roman; Marcon, Magda; Scholz, Bernhard; Calcagni, Maurizio

    2014-01-01

    The aim of this study was to evaluate a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography (FDCT) of scaphoid fixation screws. FDCT has gained interest in imaging small anatomic structures of the appendicular skeleton. Angiographic C-arm systems with flat detectors allow fluoroscopy and FDCT imaging in a one-stop procedure emphasizing their role as an ideal intraoperative imaging tool. However, FDCT imaging can be significantly impaired by artefacts induced by fixation screws. Following ethical board approval, commercially available scaphoid fixation screws were inserted into six cadaveric specimens in order to fix artificially induced scaphoid fractures. FDCT images corrected with the algorithm were compared to uncorrected images both quantitatively and qualitatively by two independent radiologists in terms of artefacts, screw contour, fracture line visibility, bone visibility, and soft tissue definition. Normal distribution of variables was evaluated using the Kolmogorov-Smirnov test. In case of normal distribution, quantitative variables were compared using paired Student's t tests. The Wilcoxon signed-rank test was used for quantitative variables without normal distribution and all qualitative variables. A p value of < 0.05 was considered to indicate statistically significant differences. Metal artefacts were significantly reduced by the correction algorithm (p < 0.001), and the fracture line was more clearly defined (p < 0.01). The inter-observer reliability was ''almost perfect'' (intra-class correlation coefficient 0.85, p < 0.001). The prototype correction algorithm in FDCT for metal artefacts induced by scaphoid fixation screws may facilitate intra- and postoperative follow-up imaging. (orig.)

  16. C-arm cone beam computed tomography needle path overlay for fluoroscopic guided vertebroplasty.

    Science.gov (United States)

    Tam, Alda L; Mohamed, Ashraf; Pfister, Marcus; Chinndurai, Ponraj; Rohm, Esther; Hall, Andrew F; Wallace, Michael J

    2010-05-01

    Retrospective review. To report our early clinical experience using C-arm cone beam computed tomography (C-arm CBCT) with fluoroscopic overlay for needle guidance during vertebroplasty. C-arm CBCT is advanced three-dimensional (3-D) imaging technology that is currently available on state-of-the-art flat panel based angiography systems. The imaging information provided by C-arm CBCT allows for the acquisition and reconstruction of "CT-like" images in flat panel based angiography/interventional suites. As part of the evolution of this technology, enhancements allowing the overlay of cross-sectional imaging information can now be integrated with real time fluoroscopy. We report our early clinical experience with C-arm CBCT with fluoroscopic overlay for needle guidance during vertebroplasty. This is a retrospective review of 10 consecutive oncology patients who underwent vertebroplasty of 13 vertebral levels using C-arm CBCT with fluoroscopic overlay for needle guidance from November 2007 to December 2008. Procedural data including vertebral level, approach (transpedicular vs. extrapedicular), access (bilateral vs. unilateral) and complications were recorded. Technical success with the overlay technology was assessed based on accuracy which consisted of 4 measured parameters: distance from target to needle tip, distance from planned path to needle tip, distance from midline to needle tip, and distance from the anterior 1/3 of the vertebral body to needle tip. Success within each parameter required that the distance between the needle tip and parameter being evaluated be no more than 5 mm on multiplanar CBCT or fluoroscopy. Imaging data for 12 vertebral levels was available for review. All vertebral levels were treated using unilateral access and 9 levels were treated with an extrapedicular approach. Technical success rates were 92% for both distance from planned path and distance from midline to final needle tip, 100% when distance from needle tip to the anterior 1

  17. Flat-Panel Computed Tomography (DYNA-CT) in Neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke.

    Science.gov (United States)

    Doerfler, A; Gölitz, P; Engelhorn, T; Kloska, S; Struffert, T

    2015-10-01

    Originally aimed at improving standard radiography by providing higher absorption efficiency and a wider dynamic range, flat-panel detector technology has meanwhile got widely accepted in the neuroradiological community. Especially flat-panel detector computed tomography (FD-CT) using rotational C-arm mounted flat-panel detector technology is capable of volumetric imaging with a high spatial resolution. By providing CT-like images of the brain within the angio suite, FD-CT is able to rapidly visualize hemorrhage and may thus improve complication management without the need of patient transfer. As "Angiographic CT" FD-CT may be helpful during many diagnostic and neurointerventional procedures and for noninvasive monitoring and follow-up. In addition, spinal interventions and high-resolution imaging of the temporal bone might also benefit from FD-CT. Finally, using novel dynamic perfusion and angiographic protocols, FD-CT may provide functional information on brain perfusion and vasculature with the potential to replace standard imaging in selected acute stroke patients.

  18. Portable low-cost flat panel detectors for real-time digital radiography

    International Nuclear Information System (INIS)

    Iovea, Mihai; Neagu, Marian; Stefanescu, Bogdan; Mateiasi, Gabriela; Porosnicu, Ioana; Angheluta, Elena

    2015-01-01

    The X-ray inspection is one of the most common used non-destructive testing methods in industry applications, but for the portable X-ray digital solution are not so many accessible, low-cost and versatile detection devices. The efficiency of a non-destructive X-ray portable device is represented by the quality of digital images, by its low acquisition time combined with a high resolution, in condition of low noise and at an affordable cost. The paper presents two X-ray portable imaging systems developed by us, suitable also for aerospace NDT applications, which are also very versatile for being easily adapted for other fields that requires mobile solutions. The first device described in the paper represent a portable large-size (210 mm X 550 mm) and high-resolution (27/54 microns) flat panel detector based on linear translation of a X-Ray TDI detector, destined for various components/parts real-time transmission measurements. The second system it is also a flat panel detectors, with a size of 510 mm X 610 mm, with the detector size from 0.2 mm until 1.5 mm, which can operate by applying the dual-energy method, very useful for discriminating materials by evaluating their Atomic effective number. The high resolution and low-cost of this flat-panels widens their applicability by covering large requirements, from identifying unwanted materials within a structure until detection of very thin cracks in complex components.

  19. Portable low-cost flat panel detectors for real-time digital radiography

    Energy Technology Data Exchange (ETDEWEB)

    Iovea, Mihai; Neagu, Marian; Stefanescu, Bogdan; Mateiasi, Gabriela; Porosnicu, Ioana; Angheluta, Elena [Accent Pro 2000 S.R.L., Bucharest (Romania)

    2015-07-01

    The X-ray inspection is one of the most common used non-destructive testing methods in industry applications, but for the portable X-ray digital solution are not so many accessible, low-cost and versatile detection devices. The efficiency of a non-destructive X-ray portable device is represented by the quality of digital images, by its low acquisition time combined with a high resolution, in condition of low noise and at an affordable cost. The paper presents two X-ray portable imaging systems developed by us, suitable also for aerospace NDT applications, which are also very versatile for being easily adapted for other fields that requires mobile solutions. The first device described in the paper represent a portable large-size (210 mm X 550 mm) and high-resolution (27/54 microns) flat panel detector based on linear translation of a X-Ray TDI detector, destined for various components/parts real-time transmission measurements. The second system it is also a flat panel detectors, with a size of 510 mm X 610 mm, with the detector size from 0.2 mm until 1.5 mm, which can operate by applying the dual-energy method, very useful for discriminating materials by evaluating their Atomic effective number. The high resolution and low-cost of this flat-panels widens their applicability by covering large requirements, from identifying unwanted materials within a structure until detection of very thin cracks in complex components.

  20. Metallic artifacts from internal scaphoid fracture fixation screws: comparison between C-arm flat-panel, cone-beam, and multidetector computed tomography.

    Science.gov (United States)

    Finkenstaedt, Tim; Morsbach, Fabian; Calcagni, Maurizio; Vich, Magdalena; Pfirrmann, Christian W A; Alkadhi, Hatem; Runge, Val M; Andreisek, Gustav; Guggenberger, Roman

    2014-08-01

    The aim of this study was to compare image quality and extent of artifacts from scaphoid fracture fixation screws using different computed tomography (CT) modalities and radiation dose protocols. Imaging of 6 cadaveric wrists with artificial scaphoid fractures and different fixation screws was performed in 2 screw positions (45° and 90° orientation in relation to the x/y-axis) using multidetector CT (MDCT) and 2 flat-panel CT modalities, C-arm flat-panel CT (FPCT) and cone-beam CT (CBCT), the latter 2 with low and standard radiation dose protocols. Mean cartilage attenuation and metal artifact-induced absolute Hounsfield unit changes (= artifact extent) were measured. Two independent radiologists evaluated different image quality criteria using a 5-point Likert-scale. Interreader agreements (Cohen κ) were calculated. Mean absolute Hounsfield unit changes and quality ratings were compared using Friedman and Wilcoxon signed-rank tests. Artifact extent was significantly smaller for MDCT and standard-dose FPCT compared with CBCT low- and standard-dose acquisitions (all P 0.05). Both MDCT and FPCT standard-dose protocols showed equal ratings for screw bone interface, fracture line, and trabecular bone evaluation (P = 0.06, 0.2, and 0.2, respectively) and performed significantly better than FPCT low- and CBCT low- and standard-dose acquisitions (all P < 0.05). Good interreader agreement was found for image quality comparisons (Cohen κ = 0.76-0.78). Both MDCT and FPCT standard-dose acquisition showed comparatively less metal-induced artifacts and better overall image quality compared with FPCT low-dose and both CBCT acquisitions. Flat-panel CT may provide sufficient image quality to serve as a versatile CT alternative for postoperative imaging of internally fixated wrist fractures.

  1. Cone-beam CT with a flat-panel detector: From image science to image-guided surgery

    International Nuclear Information System (INIS)

    Siewerdsen, Jeffrey H.

    2011-01-01

    The development of large-area flat-panel X-ray detectors (FPDs) has spurred investigation in a spectrum of advanced medical imaging applications, including tomosynthesis and cone-beam CT (CBCT). Recent research has extended image quality metrics and theoretical models to such applications, providing a quantitative foundation for the assessment of imaging performance as well as a general framework for the design, optimization, and translation of such technologies to new applications. For example, cascaded systems models of the Fourier domain metrics, such as noise-equivalent quanta (NEQ), have been extended to these modalities to describe the propagation of signal and noise through the image acquisition and reconstruction chain and to quantify the factors that govern spatial resolution, image noise, and detectability. Moreover, such models have demonstrated basic agreement with human observer performance for a broad range of imaging conditions and imaging tasks. These developments in image science have formed a foundation for the knowledgeable development and translation of CBCT to new applications in image-guided interventions-for example, CBCT implemented on a mobile surgical C-arm for intraoperative 3D imaging. The ability to acquire high-quality 3D images on demand during surgical intervention overcomes conventional limitations of surgical guidance in the context of preoperative images alone. A prototype mobile C-arm developed in academic-industry partnership demonstrates CBCT with low radiation dose, sub-mm spatial resolution, and soft-tissue visibility potentially approaching that of diagnostic CT. Integration of the 3D imaging system with real-time tracking, deformable registration, endoscopic video, and 3D visualization offers a promising addition to the surgical arsenal in interventions ranging from head-and-neck/skull base surgery to spine, orthopaedic, thoracic, and abdominal surgeries. Cadaver studies show the potential for significant boosts in surgical

  2. Cone-beam CT with a flat-panel detector: From image science to image-guided surgery

    Energy Technology Data Exchange (ETDEWEB)

    Siewerdsen, Jeffrey H., E-mail: jeff.siewerdsen@jhu.edu [Department of Biomedical Engineering, Johns Hopkins University, Traylor Building, Room 718, 720 Rutland Avenue, Baltimore, MD 21205 (United States)

    2011-08-21

    The development of large-area flat-panel X-ray detectors (FPDs) has spurred investigation in a spectrum of advanced medical imaging applications, including tomosynthesis and cone-beam CT (CBCT). Recent research has extended image quality metrics and theoretical models to such applications, providing a quantitative foundation for the assessment of imaging performance as well as a general framework for the design, optimization, and translation of such technologies to new applications. For example, cascaded systems models of the Fourier domain metrics, such as noise-equivalent quanta (NEQ), have been extended to these modalities to describe the propagation of signal and noise through the image acquisition and reconstruction chain and to quantify the factors that govern spatial resolution, image noise, and detectability. Moreover, such models have demonstrated basic agreement with human observer performance for a broad range of imaging conditions and imaging tasks. These developments in image science have formed a foundation for the knowledgeable development and translation of CBCT to new applications in image-guided interventions-for example, CBCT implemented on a mobile surgical C-arm for intraoperative 3D imaging. The ability to acquire high-quality 3D images on demand during surgical intervention overcomes conventional limitations of surgical guidance in the context of preoperative images alone. A prototype mobile C-arm developed in academic-industry partnership demonstrates CBCT with low radiation dose, sub-mm spatial resolution, and soft-tissue visibility potentially approaching that of diagnostic CT. Integration of the 3D imaging system with real-time tracking, deformable registration, endoscopic video, and 3D visualization offers a promising addition to the surgical arsenal in interventions ranging from head-and-neck/skull base surgery to spine, orthopaedic, thoracic, and abdominal surgeries. Cadaver studies show the potential for significant boosts in surgical

  3. Intraoperative imaging for patient safety and QA: detection of intracranial hemorrhage using C-arm cone-beam CT

    Science.gov (United States)

    Schafer, Sebastian; Wang, Adam; Otake, Yoshito; Stayman, J. W.; Zbijewski, Wojciech; Kleinszig, Gerhard; Xia, Xuewei; Gallia, Gary L.; Siewerdsen, Jeffrey H.

    2013-03-01

    Intraoperative imaging could improve patient safety and quality assurance (QA) via the detection of subtle complications that might otherwise only be found hours after surgery. Such capability could therefore reduce morbidity and the need for additional intervention. Among the severe adverse events that could be more quickly detected by high-quality intraoperative imaging is acute intracranial hemorrhage (ICH), conventionally assessed using post-operative CT. A mobile C-arm capable of high-quality cone-beam CT (CBCT) in combination with advanced image reconstruction techniques is reported as a means of detecting ICH in the operating room. The system employs an isocentric C-arm with a flat-panel detector in dual gain mode, correction of x-ray scatter and beam-hardening, and a penalized likelihood (PL) iterative reconstruction method. Performance in ICH detection was investigated using a quantitative phantom focusing on (non-contrast-enhanced) blood-brain contrast, an anthropomorphic head phantom, and a porcine model with injection of fresh blood bolus. The visibility of ICH was characterized in terms of contrast-to-noise ratio (CNR) and qualitative evaluation of images by a neurosurgeon. Across a range of size and contrast of the ICH as well as radiation dose from the CBCT scan, the CNR was found to increase from ~2.2-3.7 for conventional filtered backprojection (FBP) to ~3.9-5.4 for PL at equivalent spatial resolution. The porcine model demonstrated superior ICH detectability for PL. The results support the role of high-quality mobile C-arm CBCT employing advanced reconstruction algorithms for detecting subtle complications in the operating room at lower radiation dose and lower cost than intraoperative CT scanners and/or fixedroom C-arms. Such capability could present a potentially valuable aid to patient safety and QA.

  4. Unequal arm space-borne gravitational wave detectors

    International Nuclear Information System (INIS)

    Larson, Shane L.; Hellings, Ronald W.; Hiscock, William A.

    2002-01-01

    Unlike ground-based interferometric gravitational wave detectors, large space-based systems will not be rigid structures. When the end stations of the laser interferometer are freely flying spacecraft, the armlengths will change due to variations in the spacecraft positions along their orbital trajectories, so the precise equality of the arms that is required in a laboratory interferometer to cancel laser phase noise is not possible. However, using a method discovered by Tinto and Armstrong, a signal can be constructed in which laser phase noise exactly cancels out, even in an unequal arm interferometer. We examine the case where the ratio of the armlengths is a variable parameter, and compute the averaged gravitational wave transfer function as a function of that parameter. Example sensitivity curve calculations are presented for the expected design parameters of the proposed LISA interferometer, comparing it to a similar instrument with one arm shortened by a factor of 100, showing how the ratio of the armlengths will affect the overall sensitivity of the instrument

  5. Tracking brachytherapy sources using emission imaging with one flat panel detector

    International Nuclear Information System (INIS)

    Song Haijun; Bowsher, James; Das, Shiva; Yin Fangfang

    2009-01-01

    This work proposes to use the radiation from brachytherapy sources to track their dwell positions in three-dimensional (3D) space. The prototype device uses a single flat panel detector and a BB tray. The BBs are arranged in a defined pattern. The shadow of the BBs on the flat panel is analyzed to derive the 3D coordinates of the illumination source, i.e., the dwell position of the brachytherapy source. A kilovoltage x-ray source located 3.3 m away was used to align the center BB with the center pixel on the flat panel detector. For a test plan of 11 dwell positions, with an Ir-192 high dose rate unit, one projection was taken for each dwell point, and locations of the BB shadows were manually identified on the projection images. The 3D coordinates for the 11 dwell positions were reconstructed based on two BBs. The distances between dwell points were compared with the expected values. The average difference was 0.07 cm with a standard deviation of 0.15 cm. With automated BB shadow recognition in the future, this technique possesses the potential of tracking the 3D trajectory and the dwell times of a brachytherapy source in real time, enabling real time source position verification.

  6. Clinical evaluation of digital angiographic system equipped with the Safire' flat-panel detector of a direct conversion type

    International Nuclear Information System (INIS)

    Miura, Yoshiaki; Miura, Yusuke; Goto, Keiichi

    2003-01-01

    This report presents a report on clinical evaluation of our newly developed flat-panel X-ray detector of a direct conversion type, designed to provide images of a resolution higher than, or at least equal to, that ensured by X-ray photographic films, in clinical digital X-ray cinematography. This new detector was named 'Safire' the acronym of 'Shimadzu advanced flat imaging receptor', emphasizing its high technological level, such as the capability to ensure high quality of images. The clinical evaluation of Shimadzu DIGITEX Premier digital angiography system, equipped with this new flat-panel X-ray detector of a direct conversion type, has been started in March, 2003, at the Kokura Memorial Hospital in Kyushu, Japan. (author)

  7. Radiation dose reduction using a CdZnTe-based computed tomography system: Comparison to flat-panel detectors

    Energy Technology Data Exchange (ETDEWEB)

    Le, Huy Q.; Ducote, Justin L.; Molloi, Sabee [Department of Radiological Sciences, University of California, Irvine, California 92697 (United States)

    2010-03-15

    Purpose: Although x-ray projection mammography has been very effective in early detection of breast cancer, its utility is reduced in the detection of small lesions that are occult or in dense breasts. One drawback is that the inherent superposition of parenchymal structures makes visualization of small lesions difficult. Breast computed tomography using flat-panel detectors has been developed to address this limitation by producing three-dimensional data while at the same time providing more comfort to the patients by eliminating breast compression. Flat panels are charge integrating detectors and therefore lack energy resolution capability. Recent advances in solid state semiconductor x-ray detector materials and associated electronics allow the investigation of x-ray imaging systems that use a photon counting and energy discriminating detector, which is the subject of this article. Methods: A small field-of-view computed tomography (CT) system that uses CdZnTe (CZT) photon counting detector was compared to one that uses a flat-panel detector for different imaging tasks in breast imaging. The benefits afforded by the CZT detector in the energy weighting modes were investigated. Two types of energy weighting methods were studied: Projection based and image based. Simulation and phantom studies were performed with a 2.5 cm polymethyl methacrylate (PMMA) cylinder filled with iodine and calcium contrast objects. Simulation was also performed on a 10 cm breast specimen. Results: The contrast-to-noise ratio improvements as compared to flat-panel detectors were 1.30 and 1.28 (projection based) and 1.35 and 1.25 (image based) for iodine over PMMA and hydroxylapatite over PMMA, respectively. Corresponding simulation values were 1.81 and 1.48 (projection based) and 1.85 and 1.48 (image based). Dose reductions using the CZT detector were 52.05% and 49.45% for iodine and hydroxyapatite imaging, respectively. Image-based weighting was also found to have the least beam

  8. Radiation dose reduction using a CdZnTe-based computed tomography system: Comparison to flat-panel detectors

    International Nuclear Information System (INIS)

    Le, Huy Q.; Ducote, Justin L.; Molloi, Sabee

    2010-01-01

    Purpose: Although x-ray projection mammography has been very effective in early detection of breast cancer, its utility is reduced in the detection of small lesions that are occult or in dense breasts. One drawback is that the inherent superposition of parenchymal structures makes visualization of small lesions difficult. Breast computed tomography using flat-panel detectors has been developed to address this limitation by producing three-dimensional data while at the same time providing more comfort to the patients by eliminating breast compression. Flat panels are charge integrating detectors and therefore lack energy resolution capability. Recent advances in solid state semiconductor x-ray detector materials and associated electronics allow the investigation of x-ray imaging systems that use a photon counting and energy discriminating detector, which is the subject of this article. Methods: A small field-of-view computed tomography (CT) system that uses CdZnTe (CZT) photon counting detector was compared to one that uses a flat-panel detector for different imaging tasks in breast imaging. The benefits afforded by the CZT detector in the energy weighting modes were investigated. Two types of energy weighting methods were studied: Projection based and image based. Simulation and phantom studies were performed with a 2.5 cm polymethyl methacrylate (PMMA) cylinder filled with iodine and calcium contrast objects. Simulation was also performed on a 10 cm breast specimen. Results: The contrast-to-noise ratio improvements as compared to flat-panel detectors were 1.30 and 1.28 (projection based) and 1.35 and 1.25 (image based) for iodine over PMMA and hydroxylapatite over PMMA, respectively. Corresponding simulation values were 1.81 and 1.48 (projection based) and 1.85 and 1.48 (image based). Dose reductions using the CZT detector were 52.05% and 49.45% for iodine and hydroxyapatite imaging, respectively. Image-based weighting was also found to have the least beam

  9. Understanding the scatter radiation distribution during C-arm CT examination. A body phantom study

    International Nuclear Information System (INIS)

    Norimasa, Toshiyo; Kakimi, Akihiko; Takao, Yoshinori; Sasaki, Shohei; Katayama, Yutaka; Himoto, Daisuke; Izuta, Shinichiro; Ichida, Takao

    2016-01-01

    The purpose of this study was to understand the scatter radiation distribution during C-arm CT examination in the interventional radiography (IVR) room to show the escaped area and the radiation protective method. The C-arm rotates 200deg in 5 s. The tube voltage was 90 kV, and the entrance dose to the detector was 0.36 μGy/frame during C-arm CT examination. The scattered doses were measured each 50 cm from the isocenter like a grid pattern. The heights of the measurement were 50, 100, and 150 cm from the floor. The maximum scattered doses were 38.23 ± 0.60 μGy at 50 cm, 43.86 ± 20 μGy at 100 cm, and 25.78 ± 0.37 μGy at 150 cm. The scatter radiation distribution at 100 cm was the highest scattered dose. The operator should protect their reproductive gland, thyroid, and lens. The scattered dose was low behind the C-arm body and the bed, so they will be able to become the escaped area for staff. (author)

  10. Clinical evaluation of digital angiographic system equipped with the Safire' flat-panel detector of a direct conversion type

    Energy Technology Data Exchange (ETDEWEB)

    Miura, Yoshiaki; Miura, Yusuke; Goto, Keiichi [Shimadzu Corporation, Medical Systems Division, Research and Development, Kyoto (JP)] [and others

    2003-06-01

    This report presents a report on clinical evaluation of our newly developed flat-panel X-ray detector of a direct conversion type, designed to provide images of a resolution higher than, or at least equal to, that ensured by X-ray photographic films, in clinical digital X-ray cinematography. This new detector was named 'Safire' the acronym of 'Shimadzu advanced flat imaging receptor', emphasizing its high technological level, such as the capability to ensure high quality of images. The clinical evaluation of Shimadzu DIGITEX Premier digital angiography system, equipped with this new flat-panel X-ray detector of a direct conversion type, has been started in March, 2003, at the Kokura Memorial Hospital in Kyushu, Japan. (author)

  11. Automatic analysis of quality of images from X-ray digital flat detectors

    International Nuclear Information System (INIS)

    Le Meur, Y.

    2009-04-01

    Since last decade, medical imaging has grown up with the development of new digital imaging techniques. In the field of X-ray radiography, new detectors replace progressively older techniques, based on film or x-ray intensifiers. These digital detectors offer a higher sensibility and reduced overall dimensions. This work has been prepared with Trixell, the world leading company in flat detectors for medical radiography. It deals with quality control on digital images stemming from these detectors. High quality standards of medical imaging impose a close analysis of the defects that can appear on the images. This work describes a complete process for quality analysis of such images. A particular focus is given on the detection task of the defects, thanks to methods well adapted to our context of spatially correlated defects in noise background. (author)

  12. A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance.

    Science.gov (United States)

    Hua, Chiaho; Yao, Weiguang; Kidani, Takao; Tomida, Kazuo; Ozawa, Saori; Nishimura, Takenori; Fujisawa, Tatsuya; Shinagawa, Ryousuke; Merchant, Thomas E

    2017-11-01

    A ceiling-mounted robotic C-arm cone beam CT (CBCT) system was developed for use with a 190° proton gantry system and a 6-degree-of-freedom robotic patient positioner. We report on the mechanical design, system accuracy, image quality, image guidance accuracy, imaging dose, workflow, safety and collision-avoidance. The robotic CBCT system couples a rotating C-ring to the C-arm concentrically with a kV X-ray tube and a flat-panel imager mounted to the C-ring. CBCT images are acquired with flex correction and maximally 360° rotation for a 53 cm field of view. The system was designed for clinical use with three imaging locations. Anthropomorphic phantoms were imaged to evaluate the image guidance accuracy. The position accuracy and repeatability of the robotic C-arm was high (robotic CBCT system provides high-accuracy volumetric image guidance for proton therapy. Advances in knowledge: Ceiling-mounted robotic CBCT provides a viable option than CT on-rails for partial gantry and fixed-beam proton systems with the added advantage of acquiring images at the treatment isocentre.

  13. Resolution requirements for monitor viewing of digital flat-panel detector radiographs: a contrast detail analysis

    International Nuclear Information System (INIS)

    Peer, Siegfried; Giacomuzzi, Salvatore M.; Peer, Regina; Gassner, Eva; Steingruber, Iris; Jaschke, Werner

    2003-01-01

    With the introduction of digital flat-panel detector systems into clinical practice, the still unresolved question of resolution requirements for picture archiving communication system (PACS) workstation monitors has gained new momentum. This contrast detail analysis was thus performed to define the differences in observer performance in the detection of small low-contrast objects on clinical 1K and 2K monitor workstations. Images of the CDRAD 2.0 phantom were acquired at varying exposures on an indirect-type digital flat-panel detector. Three observers evaluated a total of 15 images each with respect to the threshold contrast for each detail size. The numbers of correctly identified objects were determined for all image subsets. No significant difference in the correct detection ratio was detected among the observers; however, the difference between the two types of workstations (1K vs 2K monitors) despite less than 3% was significant at a 95% confidence level. Slight but statistically significant differences exist in the detection of low-contrast nodular details visualized on 1K- and 2K-monitor workstations. Further work is needed to see if this result holds true also for comparison of clinical flat-panel detector images and may, for example, exert an influence on the diagnostic accuracy of chest X-ray readings. (orig.)

  14. Reduction of ring artifacts in CBCT: Detection and correction of pixel gain variations in flat panel detectors

    International Nuclear Information System (INIS)

    Altunbas, Cem; Lai, Chao-Jen; Zhong, Yuncheng; Shaw, Chris C.

    2014-01-01

    Purpose: In using flat panel detectors (FPD) for cone beam computed tomography (CBCT), pixel gain variations may lead to structured nonuniformities in projections and ring artifacts in CBCT images. Such gain variations can be caused by change in detector entrance exposure levels or beam hardening, and they are not accounted by conventional flat field correction methods. In this work, the authors presented a method to identify isolated pixel clusters that exhibit gain variations and proposed a pixel gain correction (PGC) method to suppress both beam hardening and exposure level dependent gain variations. Methods: To modulate both beam spectrum and entrance exposure, flood field FPD projections were acquired using beam filters with varying thicknesses. “Ideal” pixel values were estimated by performing polynomial fits in both raw and flat field corrected projections. Residuals were calculated by taking the difference between measured and ideal pixel values to identify clustered image and FPD artifacts in flat field corrected and raw images, respectively. To correct clustered image artifacts, the ratio of ideal to measured pixel values in filtered images were utilized as pixel-specific gain correction factors, referred as PGC method, and they were tabulated as a function of pixel value in a look-up table. Results: 0.035% of detector pixels lead to clustered image artifacts in flat field corrected projections, where 80% of these pixels were traced back and linked to artifacts in the FPD. The performance of PGC method was tested in variety of imaging conditions and phantoms. The PGC method reduced clustered image artifacts and fixed pattern noise in projections, and ring artifacts in CBCT images. Conclusions: Clustered projection image artifacts that lead to ring artifacts in CBCT can be better identified with our artifact detection approach. When compared to the conventional flat field correction method, the proposed PGC method enables characterization of nonlinear

  15. Contrast-detail analysis of three flat panel detectors for digital radiography

    International Nuclear Information System (INIS)

    Borasi, Giovanni; Samei, Ehsan; Bertolini, Marco; Nitrosi, Andrea; Tassoni, Davide

    2006-01-01

    In this paper we performed a contrast detail analysis of three commercially available flat panel detectors, two based on the indirect detection mechanism (GE Revolution XQ/i, system A, and Trixell/Philips Pixium 4600, system B) and one based on the direct detection mechanism (Hologic DirectRay DR 1000, system C). The experiment was conducted using standard x-ray radiation quality and a widely used contrast-detail phantom. Images were evaluated using a four alternative forced choice paradigm on a diagnostic-quality softcopy monitor. At the low and intermediate exposures, systems A and B gave equivalent performances. At the high dose levels, system A performed better than system B in the entire range of target sizes, even though the pixel size of system A was about 40% larger than that of system B. At all the dose levels, the performances of the system C (direct system) were lower than those of system A and B (indirect systems). Theoretical analyses based on the Perception Statistical Model gave similar predicted SNR T values corresponding to an observer efficiency of about 0.08 for systems A and B and 0.05 for system C

  16. Feasibility study of flexible flat-panel X-ray detectors for digital radiography

    International Nuclear Information System (INIS)

    Joe, Ok La; Yun, Seung Man; Kim, Ho Kyung

    2010-01-01

    Flexible flat-panel detectors (FPDs), which utilize both organic photodiode (OPD) and organic thin-film transistor (OTFT) technologies, are recently concerned in digital radiography. The flexible FPD has several potential advantages, such as high accessibility to patient, avoidance of geometrical burr due to the oblique angle incidence of X-ray, great reduction in manufacturing cost due to jet-printing. At once, The OPD/OTFT arrays were fabricated by jet-printing techniques, mechanical robustness due to plastic substrates, and so on. In this study, we have investigated the feasibility of flexible FPD by comparing theoretical detective quantum efficiency (DQE) with that of the conventional amorphous silicon-based FPD. We chose copper phthalocyanine-fullerene (CuPc-C60) organic materials for the construction of the flexible FPD. DQE was calculated by the linear-systems transfer theory

  17. Digital chest radiography with an amorphous silicon flat-panel-detector versus a storage-phosphor system: comparison of soft-copy images

    International Nuclear Information System (INIS)

    Lee, Hyun Ju; Im, Jung Gi; Goo, Jin Mo; Lee, Chang Hyun

    2006-01-01

    We compared the soft-copy images produced by an amorphous silicon flat-panel-detector system with the images produced by a storage-phosphor radiography system for their ability to visualize anatomic regions of the chest. Two chest radiologists independently analyzed 234 posteroanterior chest radiographs obtained from 78 patients on high-resolution liquid crystal display monitors (2560 x 2048 x 8 bits). In each patient, one radiograph was obtained with a storage-phosphor system, and two radiographs were obtained via amorphous silicon flat-panel-detector radiography with and without spatial frequency filtering. After randomizing the 234 images, the interpreters rated the visibility and radiographic quality of 11 different anatomic regions. Each image was ranked on a five-point scale (1 = not visualized, 2 = poor visualization, 3 = fair visualization, 4 = good visualization, and 5 = excellent visualization). The statistical difference between each system was determined using the Wilcoxon's signed rank test. The visibility of three anatomic regions (hilum, heart border and ribs), as determined by the chest radiologist with 14 years experience (ρ < 0.05) and the visibility of the thoracic spine, as determined by the chest radiologist with 8 years experience (ρ = 0.036), on the amorphous silicon flat-panel-detector radiography prior to spatial frequency filtering were significantly superior to that on the storage-phosphor radiography. The visibility of 11 anatomic regions, as determined by the chest radiologist with 14 years experience (ρ < 0.0001) and the visibility of five anatomic regions (unobscured lung, rib, proximal airway, thoracic spine and overall appearance), as determined by the chest radiologist with 8 years experience (ρ < 0.05), on the amorphous silicon flat-panel-detector radiography after spatial frequency filtering were significantly superior to that on the storage-phosphor radiography. The amorphous silicon flat-panel-detector system depicted the

  18. Solid-state, flat-panel, digital radiography detectors and their physical imaging characteristics

    Energy Technology Data Exchange (ETDEWEB)

    Cowen, A.R. [LXi Research, Academic Unit of Medical Physics, University of Leeds, West Yorkshire (United Kingdom)], E-mail: a.r.cowen@leeds.ac.uk; Kengyelics, S.M.; Davies, A.G. [LXi Research, Academic Unit of Medical Physics, University of Leeds, West Yorkshire (United Kingdom)

    2008-05-15

    Solid-state, digital radiography (DR) detectors, designed specifically for standard projection radiography, emerged just before the turn of the millennium. This new generation of digital image detector comprises a thin layer of x-ray absorptive material combined with an electronic active matrix array fabricated in a thin film of hydrogenated amorphous silicon (a-Si:H). DR detectors can offer both efficient (low-dose) x-ray image acquisition plus on-line readout of the latent image as electronic data. To date, solid-state, flat-panel, DR detectors have come in two principal designs, the indirect-conversion (x-ray scintillator-based) and the direct-conversion (x-ray photoconductor-based) types. This review describes the underlying principles and enabling technologies exploited by these designs of detector, and evaluates their physical imaging characteristics, comparing performance both against each other and computed radiography (CR). In standard projection radiography indirect conversion DR detectors currently offer superior physical image quality and dose efficiency compared with direct conversion DR and modern point-scan CR. These conclusions have been confirmed in the findings of clinical evaluations of DR detectors. Future trends in solid-state DR detector technologies are also briefly considered. Salient innovations include WiFi-enabled, portable DR detectors, improvements in x-ray absorber layers and developments in alternative electronic media to a-Si:H.

  19. C-arm Cone Beam Computed Tomographic Needle Path Overlay for Fluoroscopic-Guided Placement of Translumbar Central Venous Catheters

    International Nuclear Information System (INIS)

    Tam, Alda; Mohamed, Ashraf; Pfister, Marcus; Rohm, Esther; Wallace, Michael J.

    2009-01-01

    C-arm cone beam computed tomography is an advanced 3D imaging technology that is currently available on state-of-the-art flat-panel-based angiography systems. The overlay of cross-sectional imaging information can now be integrated with real-time fluoroscopy. This overlay technology was used to guide the placement of three percutaneous translumbar inferior vena cava catheters.

  20. Evaluation the image obtained from X-ray flat-panel detectors utilizing a polycrystalline CdZnTe film as the conversion layer

    International Nuclear Information System (INIS)

    Tokuda, S.; Kishihara, H.; Kaino, M.; Sato, T.

    2006-01-01

    We can expect that fluoroscopic images with a high sensitivity and excellent detective efficiency can be obtained by using a semiconductor with a small W factor for the conversion layer of X-ray flat-panel detectors, which have experienced a rapid gain inpopularity for medical and non-destructive industrial inspection uses in recent years. We believe that polycrystalline CdZnTe film formed by the closed spaced sublimation (CSS) method is a promising conversion material for next-generation high efficiency X-ray flat-panel detectors, and have previously reported the results of feasibility studies. In this paper, we present an overview of X-ray flat-panel detectors and the features of CdZnTe film, then we describe the CSS method of deposition and evaluation of the physical characteristics of CdZnTe film, and finally we present the results of our fabrication and testing of proto-type detectors utilizing CdZnTe film. (author)

  1. Digital chest radiography: flat-panel detectors or conventional radiographs

    International Nuclear Information System (INIS)

    Schaefer-Prokop, C.; Uffmann, M.; Sailer, J.; Kabalan, N.; Herold, C.; Prokop, M.

    2003-01-01

    Flat panel detectors are characterized by improved handling and increased dose efficiency. This allows for increasing of work flow efficiency and for reducing the exposure dose by about 50% compared to current systems with a sensitivity of 400. Whether the increased dose efficiency should be used to reduce acquisition dose or to increase image quality in the chest, will be shown by further clinical experience and will be also determined by the subjective preference of the radiologists. The decreased level of image noise opens new perspectives for image processing that way that elaborated multifrequency processing allows for optimizing the display of very small and low contrast structures that was so far limited by overlying image noise. Specialized applications of dual energy subtraction and temporal subtraction will also profit by the new detector technology and will be further driven forward in context with applications such as computed assisted diagnosis even though this is currently not yet broadly applied. Storage phosphor radiography still represents an important alternative technique based on its larger flexibility with respect to equipment configuration, its broader application options in intensive care and emergency radiology and due to economic reasons. These facts are further underlined by the fact that image quality also in storage phosphor radiography could be constantly increased by improving detector technology and image processing and consequently has a high standard. (orig.) [de

  2. Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: visibility of simulated microcalcifications.

    Science.gov (United States)

    Shen, Youtao; Zhong, Yuncheng; Lai, Chao-Jen; Wang, Tianpeng; Shaw, Chris C

    2013-10-01

    To measure and investigate the improvement of microcalcification (MC) visibility in cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS/CsI flat panel detector (Dexela 2923, Perkin Elmer). Aluminum wires and calcium carbonate grains of various sizes were embedded in a paraffin cylinder to simulate imaging of calcifications in a breast. Phantoms were imaged with a benchtop experimental cone beam CT system at various exposure levels. In addition to the Dexela detector, a high pitch (50 μm), thin (150 μm) scintillator CMOS/CsI flat panel detector (C7921CA-09, Hamamatsu Corporation, Hamamatsu City, Japan) and a widely used low pitch (194 μm), thick (600 μm) scintillator aSi/CsI flat panel detector (PaxScan 4030CB, Varian Medical Systems) were also used in scanning for comparison. The images were independently reviewed by six readers (imaging physicists). The MC visibility was quantified as the fraction of visible MCs and measured as a function of the estimated mean glandular dose (MGD) level for various MC sizes and detectors. The modulation transfer functions (MTFs) and detective quantum efficiencies (DQEs) were also measured and compared for the three detectors used. The authors have demonstrated that the use of a high pitch (75 μm) CMOS detector coupled with a thick (500 μm) CsI scintillator helped make the smaller 150-160, 160-180, and 180-200 μm MC groups more visible at MGDs up to 10.8, 9, and 10.8 mGy, respectively. It also made the larger 200-212 and 212-224 μm MC groups more visible at MGDs up to 7.2 mGy. No performance improvement was observed for 224-250 μm or larger size groups. With the higher spatial resolution of the Dexela detector based system, the apparent dimensions and shapes of MCs were more accurately rendered. The results show that with the aforementioned detector, a 73% visibility could be achieved in imaging 160-180 μm MCs as compared to 28% visibility achieved by the low pitch (194 μm) aSi/CsI flat

  3. WE-EF-207-02: The Rotate-Plus-Shift C-Arm Trajectory: Theory and First Clinical Results

    International Nuclear Information System (INIS)

    Ritschl, L; Kachelriess, M; Kuntz, J

    2015-01-01

    Purpose: The proposed method enables the acquisition of a complete dataset for 3D reconstruction of C-Arm data using less than 180° rotation. Methods: Typically a C–arm cone–beam CT scan is performed using a circle–like trajectory around a region of interest. Therefore an angular range of at least 180° plus fan–angle must be covered to ensure a completely sampled data set. This fact defines some constraints on the geometry and technical specifications of a C–arm system, for example a larger C radius or a smaller C opening respectively. This is even more important for mobile C-arm devices which are typically used in surgical applications.To overcome these limitations we propose a new trajectory which requires only 180° minusfan–angle of rotation for a complete data set. The trajectory consists of three parts: A rotation of the C around a defined iso–center and two translational movements parallel to the detector plane at the begin and at the end of the rotation (rotate plus shift trajectory). This enables the acquisition of a completely sampled dataset using only 180° minus fan–angle of rotation. Results: For the evaluation of the method we show simulated and measured data. The results show, that the rotate plus shift scan yields equivalent image quality compared to the short scan which is assumed to be the gold standard for C-arm CT today. Compared to the pure rotational scan over only 165°, the rotate plus shift scan shows strong improvements in image quality. Conclusion: The proposed method makes 3D imaging using C–arms with less than 180° rotation range possible. This enables integrating full 3D functionality into a C- arm device without any loss of handling and usability for 2D imaging

  4. Developing low-dose C-arm CT imaging for temporomandibular joint (TMJ) disorder in interventional radiology

    Energy Technology Data Exchange (ETDEWEB)

    Zhu, Xiaowei; Cahill, Anne Marie [Children' s Hospital of Philadelphia, Department of Radiology, Philadelphia, PA (United States); Felice, Marc [University of Pennsylvania, Environmental Health and Radiation Safety, Philadelphia, PA (United States); Johnson, Laura [Computed Tomography Division, Siemens Healthcare Sector, Shanghai (China); Sarmiento, Marily [Siemens Medical Solutions, Angiography and X-ray Division, Hoffman Estates, IL (United States)

    2011-04-15

    Manufacturers have provided C-arm CT imaging technologies for applications in interventional radiology in recent years. However, clinical imaging protocols and radiation doses have not been well studied or reported. The purpose of this study is to develop low-dose settings for clinically acceptable CT imaging of temporomandibular joint in interventional radiology suites, using a C-arm imaging angiography system. CT scans were performed with a flat-panel digital C-arm angiographic system on a 5-year-old anthropomorphic phantom. The CTDI was determined for various rotation times, dose settings and Cu filter selections. The CTDI values were compared with those of conventional low-dose CT for the same phantom. The effectiveness of using Cu filters to reduce dose was also investigated. Images were reviewed by a senior radiologist for clinical acceptance. The manufacturer's default setting gave an equivalent CTDI of 4.8 mGy. Optimizing the dose settings and adding copper filtration reduced the radiation dose by 94%. This represents a 50% reduction from conventional CT. Use of Cu filters and low-dose settings significantly reduced radiation dose from that of standard settings. This phantom study process successfully guided the clinical implementation of low-dose studies for all ages at our institution. (orig.)

  5. Comparison of dose and image quality of a Flat-panel detector and an image intensifier

    International Nuclear Information System (INIS)

    Lazzaro, M.; Friedrich, B.Q.; Luz, R.M. da; Silva, A.M.M. da

    2016-01-01

    With the development of new technologies, have emerged new conversion methods of X-ray image, such as flat panel detectors. The aim of this work is the comparison of entrance surface air kerma (ESAK) and image quality between an image intensifier type of detector (A) and a flat panel (B). The ESAK was obtained by placing a ionization chamber under PMMA simulators of 10, 20 and 30 cm and the image quality was obtained by using the TOR "1"8FG simulator. The ESAK to the equipment A is higher when compared to the equipment B. The high contrast resolution is better for the equipment A for all thicknesses of simulators. The equipment A has low contrast resolution with a better viewing threshold for thicknesses of 10 and 20 cm, and a worse performance for 30 cm. It is concluded that the equipment B has ESAK smaller and despite having lower resolution, in almost all cases, have appropriate image quality for diagnosis. (author)

  6. Antiscatter grids in mobile C-arm cone-beam CT: Effect on image quality and dose

    Energy Technology Data Exchange (ETDEWEB)

    Schafer, S.; Stayman, J.W.; Zbijewski, W.; Schmidgunst, C.; Kleinszig, G.; Siewerdsen, J.H. [Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21202 (United States); Siemens Healthcare XP Division, Erlangen, Bavaria 91052 (Germany); Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21202 (United States) and Department of Computer Science, Johns Hopkins University, Baltimore, Maryland 21218 (United States)

    2012-01-15

    Purpose: X-ray scatter is a major detriment to image quality in cone-beam CT (CBCT). Existing geometries exhibit strong differences in scatter susceptibility with more compact geometries, e.g., dental or musculoskeletal, benefiting from antiscatter grids, whereas in more extended geometries, e.g., IGRT, grid use carries tradeoffs in image quality per unit dose. This work assesses the tradeoffs in dose and image quality for grids applied in the context of low-dose CBCT on a mobile C-arm for image-guided surgery. Methods: Studies were performed on a mobile C-arm equipped with a flat-panel detector for high-quality CBCT. Antiscatter grids of grid ratio (GR) 6:1-12:1, 40 lp/cm, were tested in ''body'' surgery, i.e., spine, using protocols for bone and soft-tissue visibility in the thoracic and abdominal spine. Studies focused on grid orientation, CT number accuracy, image noise, and contrast-to-noise ratio (CNR) in quantitative phantoms at constant dose. Results: There was no effect of grid orientation on possible gridline artifacts, given accurate angle-dependent gain calibration. Incorrect calibration was found to result in gridline shadows in the projection data that imparted high-frequency artifacts in 3D reconstructions. Increasing GR reduced errors in CT number from 31%, thorax, and 37%, abdomen, for gridless operation to 2% and 10%, respectively, with a 12:1 grid, while image noise increased by up to 70%. The CNR of high-contrast objects was largely unaffected by grids, but low-contrast soft-tissues suffered reduction in CNR, 2%-65%, across the investigated GR at constant dose. Conclusions: While grids improved CT number accuracy, soft-tissue CNR was reduced due to attenuation of primary radiation. CNR could be restored by increasing dose by factors of {approx}1.6-2.5 depending on GR, e.g., increase from 4.6 mGy for the thorax and 12.5 mGy for the abdomen without antiscatter grids to approximately 12 mGy and 30 mGy, respectively, with a high

  7. Digital radiography with a large-scale electronic flat-panel detector vs screen-film radiography: observer preference in clinical skeletal diagnostics

    International Nuclear Information System (INIS)

    Hamers, S.; Freyschmidt, J.; Neitzel, U.

    2001-01-01

    The imaging performance of a recently developed digital flat-panel detector system was compared with conventional screen-film imaging in an observer preference study. In total, 34 image pairs of various regions of the skeleton were obtained in 24 patients; 30 image pairs were included in the study. The conventional images were acquired with 250- and 400-speed screen-film combinations, using the standard technique of our department. Within hours, the digital images were obtained using identical exposure parameters. The digital system employed a large-area (43 x 43 cm) flat-panel detector based on amorphous silicon (Trixell Pixium 4600), integrated in a Bucky table. Six radiologists independently evaluated the image pairs with respect to image latitude, soft tissue rendition, rendition of the periosteal and enosteal border of cortical bone, rendition of cancellous bone and the visibility of potentially present pathological changes, using a subjective five-point scale. The digital images were rated significantly (p=0.001) better than the screen-film images with respect to soft tissue rendition and image latitude. Also the rendition of the cancellous bone and the periosteal and enosteal border of the cortical bone was rated significantly (p=0.05) better for the flat-panel detector. The visibility of pathological lesions was equivalent; only large-area sclerotic lesions (n=2) were seen superiorly on screen-film images. The new digital flat-panel detector based on amorphous silicon appears to be at least equivalent to conventional screen-film combinations for skeletal examinations, and in most respects even superior. (orig.)

  8. Performance Characteristic of a CsI(Tl) Flat Panel Detector Radiography System

    International Nuclear Information System (INIS)

    Jeong, Hoi Woun; Min, Jung Hwan; Kim, Jung Min; Park, Min Seok; Lee, Gaung Young

    2012-01-01

    The purpose of this work was to evaluate an amorphous silicon cesium iodide based indirect flat-panel detector (FPD) in terms of their modulation transfer function (MTF), Wiener spectrum (WS, or noise power spectrum, NPS), and detective quantum efficiency (DQE). Measurements were made on flat-panel detector using the International Electrotechnical Commission (IEC) defined RQA3, RQA5, RQA7, and RQA9 radiographic technique. The MTFs of the systems were measured using an edge method. The WS(NPS) of the systems were determined for a range of exposure levels by two-dimensional (2D). Fourier analysis of uniformly exposed radiographs. The DQEs were assessed from the measured MTF, WS(NPS), exposure, and estimated ideal signal-to-noise ratios. Characteristic curve in the RQA3 showed difference in the characteristic curve from RQA5, RQA7, RQA9. MTFs were not differences according to x-ray beam quality. WS(NPS) was reduced with increasing dose, and RQA 3, RQA5, RQA7, RQA9 as the order is reduced. DQE represented the best in the 1mR, RQA 3, RQA5, RQA7, RQA9 decrease in the order. The physical imaging characteristics of FPD may also differ from input beam quality. This study gives an initial motivation that the physical imaging characteristics of FPD is an important issue for the right use of digital radiography system.

  9. Performance Characteristic of a CsI(Tl) Flat Panel Detector Radiography System

    Energy Technology Data Exchange (ETDEWEB)

    Jeong, Hoi Woun [Dept. of Radiological Science, Baekseok Culture University, Seoul (Korea, Republic of); Min, Jung Hwan [Dept. of Radiological Technology, Shingu University, Seongnam (Korea, Republic of); Kim, Jung Min [Dept. of Radiological Science, Korea University, Health Science College, Seoul (Korea, Republic of); Park, Min Seok [Korea Institue of Radiological and Medical Sicences, Research Institute of Radiologycal and Medical Sciences, Seoul (Korea, Republic of); Lee, Gaung Young [National Institute of Food and Drug Safety Evaluation, Seoul (Korea, Republic of)

    2012-06-15

    The purpose of this work was to evaluate an amorphous silicon cesium iodide based indirect flat-panel detector (FPD) in terms of their modulation transfer function (MTF), Wiener spectrum (WS, or noise power spectrum, NPS), and detective quantum efficiency (DQE). Measurements were made on flat-panel detector using the International Electrotechnical Commission (IEC) defined RQA3, RQA5, RQA7, and RQA9 radiographic technique. The MTFs of the systems were measured using an edge method. The WS(NPS) of the systems were determined for a range of exposure levels by two-dimensional (2D). Fourier analysis of uniformly exposed radiographs. The DQEs were assessed from the measured MTF, WS(NPS), exposure, and estimated ideal signal-to-noise ratios. Characteristic curve in the RQA3 showed difference in the characteristic curve from RQA5, RQA7, RQA9. MTFs were not differences according to x-ray beam quality. WS(NPS) was reduced with increasing dose, and RQA 3, RQA5, RQA7, RQA9 as the order is reduced. DQE represented the best in the 1mR, RQA 3, RQA5, RQA7, RQA9 decrease in the order. The physical imaging characteristics of FPD may also differ from input beam quality. This study gives an initial motivation that the physical imaging characteristics of FPD is an important issue for the right use of digital radiography system.

  10. Practical expressions describing detective quantum efficiency in flat-panel detectors

    Science.gov (United States)

    Kim, H. K.

    2011-11-01

    In radiology, image quality excellence is a balance between system performance and patient dose, hence x-ray systems must be designed to ensure the maximum image quality is obtained for the lowest consistent dose. The concept of detective quantum efficiency (DQE) is widely used to quantify, understand, measure, and predict the performance of x-ray detectors and imaging systems. Cascaded linear-systems theory can be used to estimate DQE based on the system design parameters and this theoretical DQE can be utilized for determining the impact of various physical processes, such as secondary quantum sinks, noise aliasing, reabsorption noise, and others. However, the prediction of DQE usually requires tremendous efforts to determine each parameter consisting of the cascaded linear-systems model. In this paper, practical DQE formalisms assessing both the photoconductor- and scintillator-based flat-panel detectors under quantum-noise-limited operation are described. The developed formalisms are experimentally validated and discussed for their limits. The formalisms described in this paper would be helpful for the rapid prediction of the DQE performances of developing systems as well as the optimal design of systems.

  11. On site evaluation of three flat panel detectors for digital radiography

    International Nuclear Information System (INIS)

    Borasi, Giovanni; Nitrosi, Andrea; Ferrari, Paolo; Tassoni, Davide

    2003-01-01

    During a tender we evaluated the image performance of three commercially available active matrix flat panel imagers (AMFPI) for general radiography, one based on direct detection method (Se photoconductor) the other two on indirect detection method (CsI phosphor). Basic image quality parameters (MTF, NNPS, DQE) were evaluated with particular attention to dose and energy dependence. As it is known, presampling modulation transfer function (MTF) of selenium based detector is very high (at 70 kV, 2 cycles/mm, 2.5 μGy, about 0.80). Indirect detection panels exhibit a comparable (lower) resolution (at 70 kV, 2 cycles/mm, 2.5 μGy, MTF is about 0.34 for both the systems analyzed) and a more pronounced energy and dose dependence could also be noted in one of them. As a consequence of the very high resolution, the normalized noise power spectrum (NNPS) of the direct system is substantially flat, very similar to a white noise. Considering that the sensitive layer of all detectors is the same (0.5 mm), the relatively higher NNPS values are related to selenium absorption properties (lower Z respect to CsI:Tl) and detector inherent noise. NNPSs of the other systems, at low frequencies, are comparable but the frequency dependence is significantly different. At 70 kV, 2.5 μGy, 0.5 cycles/mm detective quantum efficiency (DQE) is about 0.35 for the direct detection system, and about the same (0.6) for the indirect ones. The combined effect of additive and multiplicative noise components makes DQE dependence on dose not monotonic. DQE present a maximum for an intermediate exposure. This complex behavior may be useful to characterize the systems in terms of the monodimensional integral over the frequency of DQE (IDQE). Both visual contrast-detail experiment and the direct evaluation of the signal-to-noise ratio confirmed, at least in a qualitative way, the system performances predicted by IDQE

  12. Development of flat panel X-ray detector utilizing a CdZnTe film as conversion layer

    International Nuclear Information System (INIS)

    Tokuda, Satoshi; Kishihara, Hiroyuki; Kaino, Masatomo; Sato, Toshiyuki

    2006-01-01

    A polycrystalline CdZnTe film formed by the CSS (closed-spaced sublimation) method is one of the most promising materials as a conversion layer of next-generation highly efficient flat-panel X-ray detectors. Therefore, we have developed a prototype of a new flat-panel X-ray detector (a sensing region of 3 inches by 3 inches) with the film and evaluated its commercial feasibility. This paper describes evaluation of the physical and imaging properties of the prototype and explains the features of the CdZnTe film and the construction, specifications, and fabrication procedures of the prototype. Also included in this paper are formation of a semiconductor thin film barrier layer by the CBD (chemical bath deposition) method and conjunction of a sensor substrate and a TFT array substrate with the bump electrodes formed by screen printing, both of which we have developed during the course of the development of the prototype. (author)

  13. TU-E-217BCD-06: Cone Beam Breast CT with a High Resolution Flat Panel Detector-Improvement of Calcification Visibility.

    Science.gov (United States)

    Shen, Y; Zhong, Y; Lai, C; Wang, T; Shaw, C

    2012-06-01

    To investigate the advantage of a high resolution flat panel detector for improving the visibility of microcalcifications (MCs) in cone beam breast CT Methods: A paraffin cylinder was used to simulate a 100% adipose breast. Calcium carbonate grains, ranging from 125-140 μm to 224 - 250 μm in size, were used to simulate the MCs. Groups of 25 same size MCs were embedded at the phantom center. The phantom was scanned with a bench-top CBCT system at various exposure levels. A 75μm pitch flat panel detector (Dexela 2923, Perkin Elmer) with 500μm thick CsI scintillator plate was used as the high resolution detector. A 194 μm pitch detector (Paxscan 4030CB, Varian Medical Systems) was used for reference. 300 projection images were acquired over 360° and reconstructed. The images were reviewed by 6 readers. The MC visibility was quantified as the fraction of visible MCs and averaged for comparison. The visibility was plotted as a function of the estimated dose level for various MC sizes and detectors. The MTFs and DQEs were measured and compared. For imaging small (200 μm and smaller) MCs, the visibility achieved with the 75μm pitch detector was found to be significantly higher than those achieved with the 194μm pitch detector. For imaging larger MCs, there was little advantage in using the 75μm pitch detector. Using the 75μm pitch detector, MCs as small as 180 μm could be imaged to achieve a visibility of 78% with an isocenter tissue dose of ∼20 mGys versus 62% achieved with the 194 μm pitch detector at the same dose level. It was found that a high pitch flat panel detector had the advantages of extending its imaging capability to higher frequencies thus helping improve the visibility when used to image small MCs. This work was supported in part by grants CA104759, CA13852 and CA124585 from NIH-NCI, a grant EB00117 from NIH-NIBIB, and a subcontract from NIST-ATP. © 2012 American Association of Physicists in Medicine.

  14. Development of the OPESCOPE mobile C-arm system

    International Nuclear Information System (INIS)

    Tsuji, Hisao; Kadowaki, Toshio; Shimizu, Yasumitsu

    1994-01-01

    A new mobile C-arm X-ray fluoroscopy system called the 'OPESCOPE' equipped with a CCD TV camera and high-definition Image Intensifier has been developed. All the cables for the I.I. and X-ray generator are held inside the C-arm to prevent interference with operation and the C-arm is locked electromagnetically. Moreover, the C-arm is spring counterbalanced in the vertical motion axis. These features enable smooth positioning and uncluttered operation. The X-ray generator uses a high-frequency inverter designed to assure noiseless operation and a compact size. With the new DFS-700 videoprocessor unit being combined, clearer, more informative images can be obtained. In addition to its inherent portability this unit can be upgraded in DSA function to allow angiographic examinations in the surgical theater. (author)

  15. Flat panel detectors - closing the (digital) gap in chest and skeletal radiology

    International Nuclear Information System (INIS)

    Reiff, Kurt J.

    1999-01-01

    In the radiological department today the majority of all X-ray procedures on chest and skeletal radiography is performed with classical film-screen-systems. Using digital luminescence radiography (DLR or CR, which stands for Computed Radiography) as a technique has shown a way to replace this 100-year-old procedure of doing general radiography work by acquiring the X-rays digitally via phosphor screens, but this approach has faced criticism from lots of radiologists world wide and therefore has not been widely accepted except in the intensive care environment. A new technology is now rising based on the use of so called flat panel X-ray (FD) detectors. Semi-conducting material detects the X-rays in digital form directly and creates an instantaneous image for display, distribution and diagnosis. This ability combined with a large field of view and -- compared to existing methods -- excellent detective quantum efficiency represents a revolutionary step for chest and skeletal radiography and will put basic X-ray-work back into the focus of radiological solutions. This paper will explain the basic technology of flat panel detectors, possible system solutions based on this new technology, aspects of the user interface influencing the system utilization and versatility as well as the possibility to redefine the patient examination process for chest and skeletal radiography. Furthermore the author discusses limitations for the first released systems, upgrades for the installed base and possible scenarios for the future. e.g. fluoroscopy or angiography application

  16. Paediatric interventional cardiology: flat detector versus image intensifier using a test object

    Science.gov (United States)

    Vano, E.; Ubeda, C.; Martinez, L. C.; Leyton, F.; Miranda, P.

    2010-12-01

    Entrance surface air kerma (ESAK) values and image quality parameters were measured and compared for two biplane angiography x-ray systems dedicated to paediatric interventional cardiology, one equipped with image intensifiers (II) and the other one with dynamic flat detectors (FDs). Polymethyl methacrylate phantoms of different thicknesses, ranging from 8 to 16 cm, and a Leeds TOR 18-FG test object were used. The parameters of the image quality evaluated were noise, signal-difference-to-noise ratio (SdNR), high contrast spatial resolution (HCSR) and three figures of merit combining entrance doses and signal-to-noise ratios or HCSR. The comparisons showed a better behaviour of the II-based system in the low contrast region over the whole interval of thicknesses. The FD-based system showed a better performance in HCSR. The FD system evaluated would need around two times more dose than the II system evaluated to reach a given value of SdNR; moreover, a better spatial resolution was measured (and perceived in conventional monitors) for the system equipped with flat detectors. According to the results of this paper, the use of dynamic FD systems does not lead to an automatic reduction in ESAK or to an automatic improvement in image quality by comparison with II systems. Any improvement also depends on the setting of the x-ray systems and it should still be possible to refine these settings for some of the dynamic FDs used in paediatric cardiology.

  17. Paediatric interventional cardiology: flat detector versus image intensifier using a test object

    Energy Technology Data Exchange (ETDEWEB)

    Vano, E [Radiology Department, Medicine School, Complutense University and San Carlos University Hospital, 28040 Madrid (Spain); Ubeda, C [Clinical Sciences Department, Faculty of the Science of Health and CIHDE, Tarapaca University, 18 de Septiembre 2222, Arica (Chile); Martinez, L C [Medical Physics and Radiation Protection Service, 12 de Octubre University Hospital, Madrid (Spain); Leyton, F [Institute of Public Health of Chile, Marathon 1000, Nunoa, Santiago (Chile); Miranda, P, E-mail: eliseov@med.ucm.e [Hemodynamic Department, Cardiovascular Service, Luis Calvo Mackenna Hospital, Avenida Antonio Varaas 360, Providencia, Santiago (Chile)

    2010-12-07

    Entrance surface air kerma (ESAK) values and image quality parameters were measured and compared for two biplane angiography x-ray systems dedicated to paediatric interventional cardiology, one equipped with image intensifiers (II) and the other one with dynamic flat detectors (FDs). Polymethyl methacrylate phantoms of different thicknesses, ranging from 8 to 16 cm, and a Leeds TOR 18-FG test object were used. The parameters of the image quality evaluated were noise, signal-difference-to-noise ratio (SdNR), high contrast spatial resolution (HCSR) and three figures of merit combining entrance doses and signal-to-noise ratios or HCSR. The comparisons showed a better behaviour of the II-based system in the low contrast region over the whole interval of thicknesses. The FD-based system showed a better performance in HCSR. The FD system evaluated would need around two times more dose than the II system evaluated to reach a given value of SdNR; moreover, a better spatial resolution was measured (and perceived in conventional monitors) for the system equipped with flat detectors. According to the results of this paper, the use of dynamic FD systems does not lead to an automatic reduction in ESAK or to an automatic improvement in image quality by comparison with II systems. Any improvement also depends on the setting of the x-ray systems and it should still be possible to refine these settings for some of the dynamic FDs used in paediatric cardiology.

  18. C-arm CT for chemo-embolization of liver tumors

    International Nuclear Information System (INIS)

    Huppert, P.E.; Firlbeck, G.; Meissner, O.A.; Wietholtz, H.

    2009-01-01

    Local efficacy of transarterial chemo-embolization (TACE) is enhanced if selective treatment is performed. Selectivity of TACE mainly depends on vascular anatomy but also on the identification and catheterization of tumor feeding arteries. Correlation of vascular territories and target tumor volume in angiographic projection images is more difficult if tumors are not hypervascularized and contrast of liver parenchyma is inhomogeneous. C-arm CT offers the option of selective perfusion imaging via tumor-feeding arteries. This allows the comparison of perfusion images and baseline cross-sectional imaging to evaluate if tumors are covered completely by local treatment and to change the catheter position if necessary. Furthermore the uptake of embolization material, such as lipiodol can be checked by C-arm CT. In a prospective study of 75 TACE of liver tumors and liver metastases we evaluated the appropriateness of 85 catheter positions ready for delivery by perfusion C-arm CT and compared the diagnostic confidence of angiography and perfusion C-arm CT in terms of judgment of correct catheter position for the planned treatment. Diagnostic confidence was improved by perfusion C-arm CT in 55% of cases and in 11 cases (13%) catheter positions were inappropriate and had to be corrected. The reasons for catheter repositioning were incomplete coverage of the target tumor by perfusion volume (mismatch) in 6 cases, inappropriate perfusion of adjacent liver parenchyma in 2 cases and non-selective tumor perfusion via collateral arteries in 3 cases. C-arm CT allowed sufficient visualization of uptake of lipiodol in all cases evaluated. The diagnostic benefit of C-arm CT increases if tumors are treated more selectively, are not strongly hypervascular, are located centrally and if the enhancement of liver parenchyma is inhomogeneous. C-arm CT causes additional working time and contrast load, which is relatively low compared to angiography. Radiation exposure of 151 μGy per C-arm

  19. Interventional C-arm tomosynthesis for vascular imaging: initial results

    Science.gov (United States)

    Langan, David A.; Claus, Bernhard E. H.; Al Assad, Omar; Trousset, Yves; Riddell, Cyril; Avignon, Gregoire; Solomon, Stephen B.; Lai, Hao; Wang, Xin

    2015-03-01

    As percutaneous endovascular procedures address more complex and broader disease states, there is an increasing need for intra-procedure 3D vascular imaging. In this paper, we investigate C-Arm 2-axis tomosynthesis ("Tomo") as an alternative to C-Arm Cone Beam Computed Tomography (CBCT) for workflow situations in which the CBCT acquisition may be inconvenient or prohibited. We report on our experience in performing tomosynthesis acquisitions with a digital angiographic imaging system (GE Healthcare Innova 4100 Angiographic Imaging System, Milwaukee, WI). During a tomo acquisition the detector and tube each orbit on a plane above and below the table respectively. The tomo orbit may be circular or elliptical, and the tomographic half-angle in our studies varied from approximately 16 to 28 degrees as a function of orbit period. The trajectory, geometric calibration, and gantry performance are presented. We overview a multi-resolution iterative reconstruction employing compressed sensing techniques to mitigate artifacts associated with incomplete data reconstructions. In this work, we focus on the reconstruction of small high contrast objects such as iodinated vasculature and interventional devices. We evaluate the overall performance of the acquisition and reconstruction through phantom acquisitions and a swine study. Both tomo and comparable CBCT acquisitions were performed during the swine study thereby enabling the use of CBCT as a reference in the evaluation of tomo vascular imaging. We close with a discussion of potential clinical applications for tomo, reflecting on the imaging and workflow results achieved.

  20. Registration of 2D C-Arm and 3D CT Images for a C-Arm Image-Assisted Navigation System for Spinal Surgery

    Directory of Open Access Journals (Sweden)

    Chih-Ju Chang

    2015-01-01

    Full Text Available C-Arm image-assisted surgical navigation system has been broadly applied to spinal surgery. However, accurate path planning on the C-Arm AP-view image is difficult. This research studies 2D-3D image registration methods to obtain the optimum transformation matrix between C-Arm and CT image frames. Through the transformation matrix, the surgical path planned on preoperative CT images can be transformed and displayed on the C-Arm images for surgical guidance. The positions of surgical instruments will also be displayed on both CT and C-Arm in the real time. Five similarity measure methods of 2D-3D image registration including Normalized Cross-Correlation, Gradient Correlation, Pattern Intensity, Gradient Difference Correlation, and Mutual Information combined with three optimization methods including Powell’s method, Downhill simplex algorithm, and genetic algorithm are applied to evaluate their performance in converge range, efficiency, and accuracy. Experimental results show that the combination of Normalized Cross-Correlation measure method with Downhill simplex algorithm obtains maximum correlation and similarity in C-Arm and Digital Reconstructed Radiograph (DRR images. Spine saw bones are used in the experiment to evaluate 2D-3D image registration accuracy. The average error in displacement is 0.22 mm. The success rate is approximately 90% and average registration time takes 16 seconds.

  1. The effect of dose reduction on image quality in digital radiography using a flat-panel detector: experimental study in rabbits

    International Nuclear Information System (INIS)

    Jung, Sung Il; Goo, Jin Mo; Lee, Hyun Ju; Moon, Woo Kyung; Lim, Kun Young; Cho, Gyung Goo; Kim, Ji Hoon; Im, Jung Gi; Choi, Jang Yong; Nam, Sang Hee

    2005-01-01

    To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector. Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 μGY for the standard dose group and 20.09 μGY for the reduced dose group). The amorphous selenium-based flat-panel detector system had a panel size of 7 x 8.5 inches, a matrix of 1280 x 1536 (pixels?), and a pixel pitch of 138 μm. Four observers evaluated the soft-copy images on a high-resolution video monitor (2560 x 2048 x 8 bits) in random order. The observers rated the visibility of 13 different anatomic structures on a 5-point scale, viz, the retrocardiac lung, subdiaphragmatic lung, heart border, diaphragmatic border, proximal airway, unobscured lung, liver border, kidney border, bowel gas, flank stripe, ribs, and vertebrae in the mediastinal and abdominal regions. Statistical significance was determined using Wilcoxon's signed rank test. There was no statistically significant difference in the visibility of the anatomic structures on digital radiography between the standard and reduced dose groups. Digital radiography using an amorphous selenium-based flat-panel detector can preserve the image quality, though the dose is reduced to 40% of the standard level

  2. The effect of dose reduction on image quality in digital radiography using a flat-panel detector: experimental study in rabbits

    Energy Technology Data Exchange (ETDEWEB)

    Jung, Sung Il; Goo, Jin Mo; Lee, Hyun Ju; Moon, Woo Kyung; Lim, Kun Young; Cho, Gyung Goo; Kim, Ji Hoon; Im, Jung Gi [Seoul National College of Medicine, Seoul (Korea, Republic of); Choi, Jang Yong; Nam, Sang Hee [Inje University, Seoul (Korea, Republic of)

    2005-07-15

    To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector. Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 {mu}GY for the standard dose group and 20.09 {mu}GY for the reduced dose group). The amorphous selenium-based flat-panel detector system had a panel size of 7 x 8.5 inches, a matrix of 1280 x 1536 (pixels?), and a pixel pitch of 138 {mu}m. Four observers evaluated the soft-copy images on a high-resolution video monitor (2560 x 2048 x 8 bits) in random order. The observers rated the visibility of 13 different anatomic structures on a 5-point scale, viz, the retrocardiac lung, subdiaphragmatic lung, heart border, diaphragmatic border, proximal airway, unobscured lung, liver border, kidney border, bowel gas, flank stripe, ribs, and vertebrae in the mediastinal and abdominal regions. Statistical significance was determined using Wilcoxon's signed rank test. There was no statistically significant difference in the visibility of the anatomic structures on digital radiography between the standard and reduced dose groups. Digital radiography using an amorphous selenium-based flat-panel detector can preserve the image quality, though the dose is reduced to 40% of the standard level.

  3. High resolution X-ray imaging of bone-implant interface by large area flat-panel detector

    International Nuclear Information System (INIS)

    Kytyr, D; Jirousek, O; Dammer, J

    2011-01-01

    The aim of the research was to investigate the cemented bone-implant interface behavior (cement layer degradation and bone-cement interface debonding) with emphasis on imaging techniques suitable to detect the early defects in the cement layer. To simulate in vivo conditions a human pelvic bone was implanted with polyurethane acetabular cup using commercial acrylic bone cement. The implanted cup was then loaded in a custom hip simulator to initiate fatigue crack propagation in the bone cement. The pelvic bone was then repetitively scanned in a micro-tomography device. Reconstructed tomography images showed failure processes that occurred in the cement layer during the first 250,000 cycles. A failure in cemented acetabular implant - debonding, crumbling and smeared cracks - has been found to be at the bone-cement interface. Use of micro-focus source and high resolution flat panel detector of large physical dimensions allowed to reconstruct the micro-structural models suitable for investigation of migration, micro-motions and consecutive loosening of the implant. The large area flat panel detector with physical dimensions 120 x 120mm with 50μm pixel size provided a superior image quality compared to clinical CT systems with 300-150μm pixel size.

  4. Neutron measurement in 12,13C+ 27Al system using CR-39 detectors and neutron rem meter

    International Nuclear Information System (INIS)

    Sahoo, G.S.; Tripathy, S.P.; Shanbhag, A.A.; Sunil, C.; Joshi, D.S.; Sarkar, P.K.

    2011-01-01

    In this work, neutron measurements carried out for the interaction of 60 and 67.5 MeV 12 C, 57.3 and 65 MeV 13 C ions with thick aluminium target by using CR-39 detectors and neutron rem meter is reported. Both the detector systems were irradiated at different angles viz. 0 deg, 30 deg, 60 deg, 90 deg with respect to the beam direction. The normalized track density measurements (tracks/cm 2 /projectile at 1m) in CR-39 detectors were correlated with the normalized dose equivalent values (μSv/projectile at 1m) obtained using the neutron rem meter. The track density was found to be more in case of 13 C than 12 C. However in all the cases, the track density per incident projectile was found to decrease as the angle with respect to beam direction increases, indicating non-isotropic nature of neutron emission. The ratio between measured dose equivalent in rem meter to the measured track densities in CR-39 detectors was found to be 2.8±0.2, which remains constant irrespective of the change in angle from beam direction as well as neutron spectrum, indicating a flat dose response of CR-39 detectors. (author)

  5. Initial experiences in clinical application of the THORAX-FD: flat-panel detector radiography in thoracic diagnosis

    International Nuclear Information System (INIS)

    Herrmann, K.A.; Staebler, A.; Bonel, H.; Kulinna, C.; Holzknecht, N.; Reiser, M.F.; Geiger, B.; Boehm, S.; Maschke, M.

    2000-01-01

    The flat-panel detector closes the gap between radiography and the digital diagnostics equipment currently in use. In addition to the dose reduction that can be expected, the availability of diagnostic information in digital form enables the user to optimize the clinical workflow and to network radiography directly with a digital archiving and communication system. (orig.)

  6. Robot arm based flat panel CT-guided electromagnetic tracked spine interventions: phantom and animal model experiments

    Energy Technology Data Exchange (ETDEWEB)

    Penzkofer, Tobias; Isfort, Peter; Bruners, Philipp; Mahnken, Andreas H. [RWTH Aachen University, Applied Medical Engineering, Helmholtz-Institute Aachen, Aachen (Germany); RWTH Aachen University, Department of Diagnostic Radiology, Aachen University Hospital, Aachen (Germany); Wiemann, Christian; Guenther, Rolf W. [RWTH Aachen University, Department of Diagnostic Radiology, Aachen University Hospital, Aachen (Germany); Kyriakou, Yiannis; Kalender, Willi A. [Friedrich-Alexander University of Erlangen-Nuremberg, Institute for Medical Physics, Erlangen (Germany); Schmitz-Rode, Thomas [RWTH Aachen University, Applied Medical Engineering, Helmholtz-Institute Aachen, Aachen (Germany)

    2010-11-15

    To evaluate accuracy and procedure times of electromagnetic tracking (EMT) in a robotic arm mounted flat panel setting using phantom and animal cadaveric models. A robotic arm mounted flat panel (RMFP) was used in combination with EMT to perform anthropomorphic phantom (n = 90) and ex vivo pig based punctures (n = 120) of lumbar facet joints (FJ, n = 120) and intervertebral discs (IVD, n = 90). Procedure accuracies and times were assessed and evaluated. FJ punctures were carried out with a spatial accuracy of 0.8 {+-} 0.9 mm (phantom) and 0.6 {+-} 0.8 mm (ex vivo) respectively. While IVD punctures showed puncture deviations of 0.6 {+-} 1.2 mm (phantom) and 0.5 {+-} 0.6 mm (ex vivo), direct and angulated phantom based punctures had accuracies of 0.8 {+-} 0.9 mm and 1.0 {+-} 1.3 mm. Planning took longer for ex vivo IVD punctures compared to phantom model interventions (39.3 {+-} 17.3 s vs. 20.8 {+-} 5.0 s, p = 0.001) and for angulated vs. direct phantom FJ punctures (19.7 {+-} 5.1 s vs. 28.6 {+-} 7.8 s, p < 0.001). Puncture times were longer for ex vivo procedures when compared to phantom model procedures in both FJ (37.9 {+-} 9.0 s vs. 23.6 {+-} 7.2 s, p = 0.001) and IVD punctures (43.9 {+-} 16.1 s vs. 31.1 {+-} 6.4 s, p = 0.026). The combination of RMFP with EMT provides an accurate method of navigation for spinal interventions such as facet joint punctures and intervertebral disc punctures. (orig.)

  7. The Development of a Hybrid-Type Radiation Detector with SiC for a Reactor Robot

    International Nuclear Information System (INIS)

    Lee, Nam Ho; Cho, Jai Wan; Kim, Seung Ho

    2005-01-01

    For a robot working in a harsh environment such as a nuclear reactor environment or a space environment, requirements of on-board radiation detectors are not the same as those for environments around human. SiC devices with the wide band-gap are less dependent on temperature than Si counterparts and the can be the better candidate for the high radiation environment. With this background, radiation performance of a commercial SiC detector in a Co-60 gamma-ray environment has been evaluated. In addition to the SiC detector, a MOSFET (Metal Oxide Semiconductor Field Effect Transistor) detector has been incorporated as a backup. With this MOSFET sensor the dosimeter can keep its radiation exposure history even with loss of power. It is not only a redundant feature but also a diverse feature. The dosimetry module can be attached to mobile robot for high radiation environment was developed. This module has both SiC diode and pMOSFET mentioned above. The monitoring program which receives the radiation information from them and gives out the alarm signal when the difference of the two values from them is over the preset level was constructed. Because both the SiC pulse-type detector and the MOSFET dosimeter are small and light weight, they can be easily accommodated on a small printcircuit board for a tight space on a robot arm or for a small spacecraft

  8. Optimal C-arm angulation during transcatheter aortic valve replacement: Accuracy of a rotational C-arm computed tomography based three dimensional heart model.

    Science.gov (United States)

    Veulemans, Verena; Mollus, Sabine; Saalbach, Axel; Pietsch, Max; Hellhammer, Katharina; Zeus, Tobias; Westenfeld, Ralf; Weese, Jürgen; Kelm, Malte; Balzer, Jan

    2016-10-26

    To investigate the accuracy of a rotational C-arm CT-based 3D heart model to predict an optimal C-arm configuration during transcatheter aortic valve replacement (TAVR). Rotational C-arm CT (RCT) under rapid ventricular pacing was performed in 57 consecutive patients with severe aortic stenosis as part of the pre-procedural cardiac catheterization. With prototype software each RCT data set was segmented using a 3D heart model. From that the line of perpendicularity curve was obtained that generates a perpendicular view of the aortic annulus according to the right-cusp rule. To evaluate the accuracy of a model-based overlay we compared model- and expert-derived aortic root diameters. For all 57 patients in the RCT cohort diameter measurements were obtained from two independent operators and were compared to the model-based measurements. The inter-observer variability was measured to be in the range of 0°-12.96° of angular C-arm displacement for two independent operators. The model-to-operator agreement was 0°-13.82°. The model-based and expert measurements of aortic root diameters evaluated at the aortic annulus ( r = 0.79, P optimal C-arm configuration, potentially simplifying current clinical workflows before and during TAVR.

  9. Prostate implant reconstruction from C-arm images with motion-compensated tomosynthesis

    International Nuclear Information System (INIS)

    Dehghan, Ehsan; Moradi, Mehdi; Wen, Xu; French, Danny; Lobo, Julio; Morris, W. James; Salcudean, Septimiu E.; Fichtinger, Gabor

    2011-01-01

    Purpose: Accurate localization of prostate implants from several C-arm images is necessary for ultrasound-fluoroscopy fusion and intraoperative dosimetry. The authors propose a computational motion compensation method for tomosynthesis-based reconstruction that enables 3D localization of prostate implants from C-arm images despite C-arm oscillation and sagging. Methods: Five C-arm images are captured by rotating the C-arm around its primary axis, while measuring its rotation angle using a protractor or the C-arm joint encoder. The C-arm images are processed to obtain binary seed-only images from which a volume of interest is reconstructed. The motion compensation algorithm, iteratively, compensates for 2D translational motion of the C-arm by maximizing the number of voxels that project on a seed projection in all of the images. This obviates the need for C-arm full pose tracking traditionally implemented using radio-opaque fiducials or external trackers. The proposed reconstruction method is tested in simulations, in a phantom study and on ten patient data sets. Results: In a phantom implanted with 136 dummy seeds, the seed detection rate was 100% with a localization error of 0.86 ± 0.44 mm (Mean ± STD) compared to CT. For patient data sets, a detection rate of 99.5% was achieved in approximately 1 min per patient. The reconstruction results for patient data sets were compared against an available matching-based reconstruction method and showed relative localization difference of 0.5 ± 0.4 mm. Conclusions: The motion compensation method can successfully compensate for large C-arm motion without using radio-opaque fiducial or external trackers. Considering the efficacy of the algorithm, its successful reconstruction rate and low computational burden, the algorithm is feasible for clinical use.

  10. Prostate implant reconstruction from C-arm images with motion-compensated tomosynthesis

    Energy Technology Data Exchange (ETDEWEB)

    Dehghan, Ehsan; Moradi, Mehdi; Wen, Xu; French, Danny; Lobo, Julio; Morris, W. James; Salcudean, Septimiu E.; Fichtinger, Gabor [School of Computing, Queen' s University, Kingston, Ontario K7L-3N6 (Canada); Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia V6T-1Z4 (Canada); Vancouver Cancer Centre, Vancouver, British Columbia V5Z-1E6 (Canada); Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia V6T-1Z4 (Canada); School of Computing, Queen' s University, Kingston, Ontario K7L-3N6 (Canada)

    2011-10-15

    Purpose: Accurate localization of prostate implants from several C-arm images is necessary for ultrasound-fluoroscopy fusion and intraoperative dosimetry. The authors propose a computational motion compensation method for tomosynthesis-based reconstruction that enables 3D localization of prostate implants from C-arm images despite C-arm oscillation and sagging. Methods: Five C-arm images are captured by rotating the C-arm around its primary axis, while measuring its rotation angle using a protractor or the C-arm joint encoder. The C-arm images are processed to obtain binary seed-only images from which a volume of interest is reconstructed. The motion compensation algorithm, iteratively, compensates for 2D translational motion of the C-arm by maximizing the number of voxels that project on a seed projection in all of the images. This obviates the need for C-arm full pose tracking traditionally implemented using radio-opaque fiducials or external trackers. The proposed reconstruction method is tested in simulations, in a phantom study and on ten patient data sets. Results: In a phantom implanted with 136 dummy seeds, the seed detection rate was 100% with a localization error of 0.86 {+-} 0.44 mm (Mean {+-} STD) compared to CT. For patient data sets, a detection rate of 99.5% was achieved in approximately 1 min per patient. The reconstruction results for patient data sets were compared against an available matching-based reconstruction method and showed relative localization difference of 0.5 {+-} 0.4 mm. Conclusions: The motion compensation method can successfully compensate for large C-arm motion without using radio-opaque fiducial or external trackers. Considering the efficacy of the algorithm, its successful reconstruction rate and low computational burden, the algorithm is feasible for clinical use.

  11. High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imaging

    Science.gov (United States)

    Abu Anas, Emran Mohammad; Kim, Jae Gon; Lee, Soo Yeol; Kamrul Hasan, Md

    2011-10-01

    The use of an x-ray flat panel detector is increasingly becoming popular in 3D cone beam volume CT machines. Due to the deficient semiconductor array manufacturing process, the cone beam projection data are often corrupted by different types of abnormalities, which cause severe ring and radiant artifacts in a cone beam reconstruction image, and as a result, the diagnostic image quality is degraded. In this paper, a novel technique is presented for the correction of error in the 2D cone beam projections due to abnormalities often observed in 2D x-ray flat panel detectors. Template images are derived from the responses of the detector pixels using their statistical properties and then an effective non-causal derivative-based detection algorithm in 2D space is presented for the detection of defective and mis-calibrated detector elements separately. An image inpainting-based 3D correction scheme is proposed for the estimation of responses of defective detector elements, and the responses of the mis-calibrated detector elements are corrected using the normalization technique. For real-time implementation, a simplification of the proposed off-line method is also suggested. Finally, the proposed algorithms are tested using different real cone beam volume CT images and the experimental results demonstrate that the proposed methods can effectively remove ring and radiant artifacts from cone beam volume CT images compared to other reported techniques in the literature.

  12. High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imaging

    International Nuclear Information System (INIS)

    Anas, Emran Mohammad Abu; Hasan, Md Kamrul; Kim, Jae Gon; Lee, Soo Yeol

    2011-01-01

    The use of an x-ray flat panel detector is increasingly becoming popular in 3D cone beam volume CT machines. Due to the deficient semiconductor array manufacturing process, the cone beam projection data are often corrupted by different types of abnormalities, which cause severe ring and radiant artifacts in a cone beam reconstruction image, and as a result, the diagnostic image quality is degraded. In this paper, a novel technique is presented for the correction of error in the 2D cone beam projections due to abnormalities often observed in 2D x-ray flat panel detectors. Template images are derived from the responses of the detector pixels using their statistical properties and then an effective non-causal derivative-based detection algorithm in 2D space is presented for the detection of defective and mis-calibrated detector elements separately. An image inpainting-based 3D correction scheme is proposed for the estimation of responses of defective detector elements, and the responses of the mis-calibrated detector elements are corrected using the normalization technique. For real-time implementation, a simplification of the proposed off-line method is also suggested. Finally, the proposed algorithms are tested using different real cone beam volume CT images and the experimental results demonstrate that the proposed methods can effectively remove ring and radiant artifacts from cone beam volume CT images compared to other reported techniques in the literature.

  13. Quantitative image quality evaluation of pixel-binning in a flat-panel detector for x-ray fluoroscopy

    International Nuclear Information System (INIS)

    Srinivas, Yogesh; Wilson, David L.

    2004-01-01

    X-ray fluoroscopy places stringent design requirements on new flat-panel (FP) detectors, requiring both low-noise electronics and high data transfer rates. Pixel-binning, wherein data from more that one detector pixel are collected simultaneously, not only lowers the data transfer rate but also increases x-ray counts and pixel signal-to-noise ratio (SNR). In this study, we quantitatively assessed image quality of image sequences from four acquisition methods; no-binning and three types of binning; in synthetic images using a clinically relevant task of detecting an extended guidewire in a four-alternative forced-choice paradigm. Binning methods were conventional data-line (D) and gate-line (G) binning, and a novel method in which alternate frames in an image sequence used D and G binning. Two detector orientations placed the data lines either parallel or perpendicular to the guide wire. At a low exposure of 0.6 μR (1.548x10 -10 C/kg) per frame, irrespective of detector orientation, D binning with its reduced electronic noise was significantly (p -10 C/kg) per frame, with data lines parallel to the guidewire, detection with D binning was significantly (p<0.1) better than G binning. However, with data lines perpendicular to the guidewire, G binning was significantly (p<0.1) better than D binning because the partial area effect was reduced. Alternate binning was the best binning method when results were averaged over both orientations, and it was as good as the best binning method at either orientation. In addition, at low and high exposures, alternate binning gave a temporally fused image with a smooth guidewire, an important image quality feature not assessed in a detection experiment. While at high exposure, detection with no binning was as good, or better, than the best binning method, it might be impractical at fluoroscopy imaging rates. A computational observer model based on signal detection theory successfully fit data and was used to predict effects of

  14. Application of C-arm computed tomography in cardiology; Kardiale Anwendung der C-Arm-Computertomographie

    Energy Technology Data Exchange (ETDEWEB)

    Rieber, J. [Klinikum der Ludwig-Maximilians-Universitaet Muenchen, Campus Innenstadt, Abteilung fuer Kardiologie, Medizinische Poliklinik, Muenchen (Germany); Rohkohl, C. [Friedrich-Alexander-Universitaet Erlangen-Nuernberg, Lehrstuhl fuer Mustererkennung, Department Informatik, Erlangen (Germany); Siemens AG, Healthcare Sector, Forchheim, Forchheim (Germany); Lauritsch, G. [Siemens AG, Healthcare Sector, Forchheim, Forchheim (Germany); Rittger, H. [Krankenhaus Coburg, Abteilung fuer Kardiologie, Coburg (Germany); Meissner, O. [Siemens AG, Healthcare Sector, Forchheim, Forchheim (Germany); Klinikum der Ludwig-Maximilians-Universitaet Muenchen, Institut fuer Klinische Radiologie, Muenchen (Germany)

    2009-09-15

    C-arm computed tomography is currently being introduced into cardiac imaging and offers the potential for three-dimensional imaging of the cardiac anatomy within the interventional environment. This detailed view is necessary to support complex interventional strategies, such as transcutaneous valve replacement, interventional therapy of atrial fibrillation, implantation of biventricular pacemakers and assessment of myocardial perfusion. Currently, the major limitation of this technology is its insufficient temporal resolution which limits the visualization of fast moving parts of the heart. (orig.) [German] Durch die Entwicklung der C-Arm-Computertomographie- (CACT-)Angiographie ist es erstmals moeglich, waehrend einer Herzkatheteruntersuchung eine detaillierte dreidimensionale Darstellung der kardialen Anatomie zu erhalten. Derartige zusaetzliche Informationen koennten die Durchfuehrung der immer komplexer werdenden Strategien der interventionellen Kardiologie wirkungsvoll unterstuetzen. Hierzu zaehlen u. a. der transkutane Klappenersatz, die interventionelle Behandlung von Vorhofflimmern, die Implantation biventrikulaerer Schrittmacher sowie die Beurteilung der Myokardperfusion. Die derzeit groesste Limitation dieser Methode ist die relativ geringe zeitliche Aufloesung, die aufgrund der Bewegung des Herzens die Anwendung dieser Technologie einschraenkt. (orig.)

  15. TU-FG-BRB-11: Design and Evaluation of a Robotic C-Arm CBCT System for Image-Guided Proton Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Hua, C; Yao, W; Farr, J; Merchant, T [St. Jude Children’s Research Hospital, Memphis, TN (United States); Kidani, T; Tomida, K; Ozawa, S; Nishimura, T; Fujusawa, T; Shinagawa, R [Hitachi, Ltd., Hitachi-shi, Ibaraki-ken (Japan)

    2016-06-15

    Purpose: To describe the design and performance of a ceiling-mounted robotic C-arm CBCT system for image-guided proton therapy. Methods: Uniquely different from traditional C-arm CBCT used in interventional radiology, the imaging system was designed to provide volumetric image guidance for patients treated on a 190-degree proton gantry system and a 6 degree-of-freedom (DOF) robotic patient positioner. The mounting of robotic arms to the ceiling rails, rather than gantry or nozzle, provides the flexibility in imaging locations (isocenter, iso+27cm in X, iso+100cm in Y) in the room and easier upgrade as technology advances. A kV X-ray tube and a 43×43cm flat panel imager were mounted to a rotating C-ring (87cm diameter), which is coupled to the C-arm concentrically. Both C-arm and the robotic arm remain stationary during imaging to maintain high position accuracy. Source-to-axis distance and source-to-imager distance are 100 and 150cm, respectively. A 14:1 focused anti-scatter grid and a bowtie filer are used for image acquisition. A unique automatic collimator device of 4 independent blades for adjusting field of view and reducing patient dose has also been developed. Results: Sub-millimeter position accuracy and repeatability of the robotic C-arm were measured with a laser tracker. High quality CBCT images for positioning can be acquired with a weighted CTDI of 3.6mGy (head in 200° full fan mode: 100kV, 20mA, 20ms, 10fps)-8.7 mGy (pelvis in 360° half fan mode: 125kV, 42mA, 20ms, 10fps). Image guidance accuracy achieved <1mm (3D vector) with automatic 3D-3D registration for anthropomorphic head and pelvis phantoms. Since November 2015, 22 proton therapy patients have undergone daily CBCT imaging for 6 DOF positioning. Conclusion: Decoupled from gantry and nozzle, this CBCT system provides a unique solution for volumetric image guidance with half/partial proton gantry systems. We demonstrated that daily CBCT can be integrated into proton therapy for pre

  16. TU-FG-BRB-11: Design and Evaluation of a Robotic C-Arm CBCT System for Image-Guided Proton Therapy

    International Nuclear Information System (INIS)

    Hua, C; Yao, W; Farr, J; Merchant, T; Kidani, T; Tomida, K; Ozawa, S; Nishimura, T; Fujusawa, T; Shinagawa, R

    2016-01-01

    Purpose: To describe the design and performance of a ceiling-mounted robotic C-arm CBCT system for image-guided proton therapy. Methods: Uniquely different from traditional C-arm CBCT used in interventional radiology, the imaging system was designed to provide volumetric image guidance for patients treated on a 190-degree proton gantry system and a 6 degree-of-freedom (DOF) robotic patient positioner. The mounting of robotic arms to the ceiling rails, rather than gantry or nozzle, provides the flexibility in imaging locations (isocenter, iso+27cm in X, iso+100cm in Y) in the room and easier upgrade as technology advances. A kV X-ray tube and a 43×43cm flat panel imager were mounted to a rotating C-ring (87cm diameter), which is coupled to the C-arm concentrically. Both C-arm and the robotic arm remain stationary during imaging to maintain high position accuracy. Source-to-axis distance and source-to-imager distance are 100 and 150cm, respectively. A 14:1 focused anti-scatter grid and a bowtie filer are used for image acquisition. A unique automatic collimator device of 4 independent blades for adjusting field of view and reducing patient dose has also been developed. Results: Sub-millimeter position accuracy and repeatability of the robotic C-arm were measured with a laser tracker. High quality CBCT images for positioning can be acquired with a weighted CTDI of 3.6mGy (head in 200° full fan mode: 100kV, 20mA, 20ms, 10fps)-8.7 mGy (pelvis in 360° half fan mode: 125kV, 42mA, 20ms, 10fps). Image guidance accuracy achieved <1mm (3D vector) with automatic 3D-3D registration for anthropomorphic head and pelvis phantoms. Since November 2015, 22 proton therapy patients have undergone daily CBCT imaging for 6 DOF positioning. Conclusion: Decoupled from gantry and nozzle, this CBCT system provides a unique solution for volumetric image guidance with half/partial proton gantry systems. We demonstrated that daily CBCT can be integrated into proton therapy for pre

  17. TU-AB-204-02: Advances in C-Arm CBCT for Cardiac Interventions

    International Nuclear Information System (INIS)

    Fahrig, R.

    2015-01-01

    , significant effort has been expended to improve the quantitative accuracy of C-arm CBCT reconstructions. The challenge is to improve image quality while providing very short turnaround between data acquisition and volume data visualization. Corrections for x-ray scatter, view aliasing and patient motion that require no more than 2 iterations keep processing time short while reducing artifact. Fast, multi-sweep acquisitions can be used to permit assessment of left ventricular function, and visualization of radiofrequency lesions created to treat arrhythmias. Workflows for each imaging goal have been developed and validated against gold standard clinical CT or histology. The challenges, opportunities, and limitations of the new functional C-arm CBCT imaging techniques will be discussed. Dr. W. Zbijewski (Johns Hopkins University) will present on the topic: Advances in CBCT for Orthopaedics and Bone Health Imaging. Cone-beam CT is particularly well suited for imaging of musculoskeletal extremities. Owing to the high spatial resolution of flat-panel detectors, CBCT can surpass conventional CT in imaging tasks involving bone visualization, quantitative analysis of subchondral trabecular structure, and visualization and monitoring of subtle fractures that are common in orthopedic radiology. A dedicated CBCT platform has been developed that offers flexibility in system design and provides not only a compact configuration with improved logistics for extremities imaging but also enables novel diagnostic capabilities such as imaging of weight-bearing lower extremities in a natural stance. The design, development and clinical performance of dedicated extremities CBCT systems will be presented. Advanced capabilities for quantitative volumetric assessment of joint space morphology, dual-energy image-based quantification of bone composition, and in-vivo analysis of bone microarchitecture will be discussed, along with emerging applications in the diagnosis of arthritis and osteoporosis and

  18. TU-AB-204-02: Advances in C-Arm CBCT for Cardiac Interventions

    Energy Technology Data Exchange (ETDEWEB)

    Fahrig, R. [Stanford University (United States)

    2015-06-15

    , significant effort has been expended to improve the quantitative accuracy of C-arm CBCT reconstructions. The challenge is to improve image quality while providing very short turnaround between data acquisition and volume data visualization. Corrections for x-ray scatter, view aliasing and patient motion that require no more than 2 iterations keep processing time short while reducing artifact. Fast, multi-sweep acquisitions can be used to permit assessment of left ventricular function, and visualization of radiofrequency lesions created to treat arrhythmias. Workflows for each imaging goal have been developed and validated against gold standard clinical CT or histology. The challenges, opportunities, and limitations of the new functional C-arm CBCT imaging techniques will be discussed. Dr. W. Zbijewski (Johns Hopkins University) will present on the topic: Advances in CBCT for Orthopaedics and Bone Health Imaging. Cone-beam CT is particularly well suited for imaging of musculoskeletal extremities. Owing to the high spatial resolution of flat-panel detectors, CBCT can surpass conventional CT in imaging tasks involving bone visualization, quantitative analysis of subchondral trabecular structure, and visualization and monitoring of subtle fractures that are common in orthopedic radiology. A dedicated CBCT platform has been developed that offers flexibility in system design and provides not only a compact configuration with improved logistics for extremities imaging but also enables novel diagnostic capabilities such as imaging of weight-bearing lower extremities in a natural stance. The design, development and clinical performance of dedicated extremities CBCT systems will be presented. Advanced capabilities for quantitative volumetric assessment of joint space morphology, dual-energy image-based quantification of bone composition, and in-vivo analysis of bone microarchitecture will be discussed, along with emerging applications in the diagnosis of arthritis and osteoporosis and

  19. A novel forward projection-based metal artifact reduction method for flat-detector computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Prell, Daniel; Kyriakou, Yiannis; Beister, Marcel; Kalender, Willi A [Institute of Medical Physics, University of Erlangen-Nuernberg, Henkestrasse 91, 91052 Erlangen (Germany)], E-mail: daniel.prell@imp.uni-erlangen.de

    2009-11-07

    Metallic implants generate streak-like artifacts in flat-detector computed tomography (FD-CT) reconstructed volumetric images. This study presents a novel method for reducing these disturbing artifacts by inserting discarded information into the original rawdata using a three-step correction procedure and working directly with each detector element. Computation times are minimized by completely implementing the correction process on graphics processing units (GPUs). First, the original volume is corrected using a three-dimensional interpolation scheme in the rawdata domain, followed by a second reconstruction. This metal artifact-reduced volume is then segmented into three materials, i.e. air, soft-tissue and bone, using a threshold-based algorithm. Subsequently, a forward projection of the obtained tissue-class model substitutes the missing or corrupted attenuation values directly for each flat detector element that contains attenuation values corresponding to metal parts, followed by a final reconstruction. Experiments using tissue-equivalent phantoms showed a significant reduction of metal artifacts (deviations of CT values after correction compared to measurements without metallic inserts reduced typically to below 20 HU, differences in image noise to below 5 HU) caused by the implants and no significant resolution losses even in areas close to the inserts. To cover a variety of different cases, cadaver measurements and clinical images in the knee, head and spine region were used to investigate the effectiveness and applicability of our method. A comparison to a three-dimensional interpolation correction showed that the new approach outperformed interpolation schemes. Correction times are minimized, and initial and corrected images are made available at almost the same time (12.7 s for the initial reconstruction, 46.2 s for the final corrected image compared to 114.1 s and 355.1 s on central processing units (CPUs))

  20. A novel forward projection-based metal artifact reduction method for flat-detector computed tomography

    International Nuclear Information System (INIS)

    Prell, Daniel; Kyriakou, Yiannis; Beister, Marcel; Kalender, Willi A

    2009-01-01

    Metallic implants generate streak-like artifacts in flat-detector computed tomography (FD-CT) reconstructed volumetric images. This study presents a novel method for reducing these disturbing artifacts by inserting discarded information into the original rawdata using a three-step correction procedure and working directly with each detector element. Computation times are minimized by completely implementing the correction process on graphics processing units (GPUs). First, the original volume is corrected using a three-dimensional interpolation scheme in the rawdata domain, followed by a second reconstruction. This metal artifact-reduced volume is then segmented into three materials, i.e. air, soft-tissue and bone, using a threshold-based algorithm. Subsequently, a forward projection of the obtained tissue-class model substitutes the missing or corrupted attenuation values directly for each flat detector element that contains attenuation values corresponding to metal parts, followed by a final reconstruction. Experiments using tissue-equivalent phantoms showed a significant reduction of metal artifacts (deviations of CT values after correction compared to measurements without metallic inserts reduced typically to below 20 HU, differences in image noise to below 5 HU) caused by the implants and no significant resolution losses even in areas close to the inserts. To cover a variety of different cases, cadaver measurements and clinical images in the knee, head and spine region were used to investigate the effectiveness and applicability of our method. A comparison to a three-dimensional interpolation correction showed that the new approach outperformed interpolation schemes. Correction times are minimized, and initial and corrected images are made available at almost the same time (12.7 s for the initial reconstruction, 46.2 s for the final corrected image compared to 114.1 s and 355.1 s on central processing units (CPUs)).

  1. Robot arm based flat panel CT-guided electromagnetic tracked spine interventions: phantom and animal model experiments

    International Nuclear Information System (INIS)

    Penzkofer, Tobias; Isfort, Peter; Bruners, Philipp; Mahnken, Andreas H.; Wiemann, Christian; Guenther, Rolf W.; Kyriakou, Yiannis; Kalender, Willi A.; Schmitz-Rode, Thomas

    2010-01-01

    To evaluate accuracy and procedure times of electromagnetic tracking (EMT) in a robotic arm mounted flat panel setting using phantom and animal cadaveric models. A robotic arm mounted flat panel (RMFP) was used in combination with EMT to perform anthropomorphic phantom (n = 90) and ex vivo pig based punctures (n = 120) of lumbar facet joints (FJ, n = 120) and intervertebral discs (IVD, n = 90). Procedure accuracies and times were assessed and evaluated. FJ punctures were carried out with a spatial accuracy of 0.8 ± 0.9 mm (phantom) and 0.6 ± 0.8 mm (ex vivo) respectively. While IVD punctures showed puncture deviations of 0.6 ± 1.2 mm (phantom) and 0.5 ± 0.6 mm (ex vivo), direct and angulated phantom based punctures had accuracies of 0.8 ± 0.9 mm and 1.0 ± 1.3 mm. Planning took longer for ex vivo IVD punctures compared to phantom model interventions (39.3 ± 17.3 s vs. 20.8 ± 5.0 s, p = 0.001) and for angulated vs. direct phantom FJ punctures (19.7 ± 5.1 s vs. 28.6 ± 7.8 s, p < 0.001). Puncture times were longer for ex vivo procedures when compared to phantom model procedures in both FJ (37.9 ± 9.0 s vs. 23.6 ± 7.2 s, p = 0.001) and IVD punctures (43.9 ± 16.1 s vs. 31.1 ± 6.4 s, p = 0.026). The combination of RMFP with EMT provides an accurate method of navigation for spinal interventions such as facet joint punctures and intervertebral disc punctures. (orig.)

  2. Heel effect adaptive flat field correction of digital x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Yu, Yongjian [X-ray Products, Varian Medical Systems Inc., Liverpool, New York 13088 (United States); Wang, Jue [Department of Mathematics, Union College, Schenectady, New York 12308 (United States)

    2013-08-15

    Purpose: Anode heel effect renders large-scale background nonuniformities in digital radiographs. Conventional offset/gain calibration is performed at mono source-to-image distance (SID), and disregards the SID-dependent characteristic of heel effect. It results in a residual nonuniform background in the corrected radiographs when the SID settings for calibration and correction differ. In this work, the authors develop a robust and efficient computational method for digital x-ray detector gain correction adapted to SID-variant heel effect, without resorting to physical filters, phantoms, complicated heel effect models, or multiple-SID calibration and interpolation.Methods: The authors present the Duo-SID projection correction method. In our approach, conventional offset/gain calibrations are performed only twice, at the minimum and maximum SIDs of the system in typical clinical use. A fast iterative separation algorithm is devised to extract the detector gain and basis heel patterns from the min/max SID calibrations. The resultant detector gain is independent of SID, while the basis heel patterns are parameterized by the min- and max-SID. The heel pattern at any SID is obtained from the min-SID basis heel pattern via projection imaging principles. The system gain desired at a specific acquisition SID is then constructed using the projected heel pattern and detector gain map.Results: The method was evaluated for flat field and anatomical phantom image corrections. It demonstrated promising improvements over interpolation and conventional gain calibration/correction methods, lowering their correction errors by approximately 70% and 80%, respectively. The separation algorithm was able to extract the detector gain and heel patterns with less than 2% error, and the Duo-SID corrected images showed perceptually appealing uniform background across the detector.Conclusions: The Duo-SID correction method has substantially improved on conventional offset/gain corrections for

  3. Heel effect adaptive flat field correction of digital x-ray detectors

    International Nuclear Information System (INIS)

    Yu, Yongjian; Wang, Jue

    2013-01-01

    Purpose: Anode heel effect renders large-scale background nonuniformities in digital radiographs. Conventional offset/gain calibration is performed at mono source-to-image distance (SID), and disregards the SID-dependent characteristic of heel effect. It results in a residual nonuniform background in the corrected radiographs when the SID settings for calibration and correction differ. In this work, the authors develop a robust and efficient computational method for digital x-ray detector gain correction adapted to SID-variant heel effect, without resorting to physical filters, phantoms, complicated heel effect models, or multiple-SID calibration and interpolation.Methods: The authors present the Duo-SID projection correction method. In our approach, conventional offset/gain calibrations are performed only twice, at the minimum and maximum SIDs of the system in typical clinical use. A fast iterative separation algorithm is devised to extract the detector gain and basis heel patterns from the min/max SID calibrations. The resultant detector gain is independent of SID, while the basis heel patterns are parameterized by the min- and max-SID. The heel pattern at any SID is obtained from the min-SID basis heel pattern via projection imaging principles. The system gain desired at a specific acquisition SID is then constructed using the projected heel pattern and detector gain map.Results: The method was evaluated for flat field and anatomical phantom image corrections. It demonstrated promising improvements over interpolation and conventional gain calibration/correction methods, lowering their correction errors by approximately 70% and 80%, respectively. The separation algorithm was able to extract the detector gain and heel patterns with less than 2% error, and the Duo-SID corrected images showed perceptually appealing uniform background across the detector.Conclusions: The Duo-SID correction method has substantially improved on conventional offset/gain corrections for

  4. Flat-panel detectors: how much better are they?

    International Nuclear Information System (INIS)

    Seibert, J.A.

    2006-01-01

    Interventional and fluoroscopic imaging procedures for pediatric patients are becoming more prevalent because of the less-invasive nature of these procedures compared to alternatives such as surgery. Flat-panel X-ray detectors (FPD) for fluoroscopy are a new technology alternative to the image intensifier/TV (II/TV) digital system that has been in use for more than two decades. Two major FPD technologies have been implemented, based on indirect conversion of X-rays to light (using an X-ray scintillator) and then to proportional charge (using a photodiode), or direct conversion of X-rays into charge (using a semiconductor material) for signal acquisition and digitization. These detectors have proved very successful for high-exposure interventional procedures but lack the image quality of the II/TV system at the lowest exposure levels common in fluoroscopy. The benefits for FPD image quality include lack of geometric distortion, little or no veiling glare, a uniform response across the field-of-view, and improved ergonomics with better patient access. Better detective quantum efficiency indicates the possibility of reducing the patient dose in accordance with ALARA principles. However, first-generation FPD devices have been implemented with less than adequate acquisition flexibility (e.g., lack of tableside controls/information, inability to easily change protocols) and the presence of residual signals from previous exposures, and additional cost of equipment and long-term maintenance have been serious impediments to purchase and implementation. Technological advances of second generation and future hybrid FPD systems should solve many current issues. The answer to the question ''how much better are they?'' is ''significantly better'', and they are certainly worth consideration for replacement or new implementation of an imaging suite for pediatric fluoroscopy. (orig.)

  5. Virtual Reality Aided Positioning of Mobile C-Arms for Image-Guided Surgery

    Directory of Open Access Journals (Sweden)

    Zhenzhou Shao

    2014-06-01

    Full Text Available For the image-guided surgery, the positioning of mobile C-arms is a key technique to take X-ray images in a desired pose for the confirmation of current surgical outcome. Unfortunately, surgeons and patient often suffer the radiation exposure due to the repeated imaging when the X-ray image is of poor quality or not captured at a good projection view. In this paper, a virtual reality (VR aided positioning method for the mobile C-arm is proposed by the alignment of 3D surface model of region of interest and preoperative anatomy, so that a reference pose of the mobile C-arm with respect to the inside anatomy can be figured out from outside view. It allows a one-time imaging from the outside view to greatly reduce the additional radiation exposure. To control the mobile C-arm to the desired pose, the mobile C-arm is modeled as a robotic arm with a movable base. Experiments were conducted to evaluate the accuracy of appearance model and precision of mobile C-arm positioning. The appearance model was reconstructed with the average error of 2.16 mm. One-time imaging of mobile C-arm was achieved, and new modeling of mobile C-arm with 8 DoFs enlarges the working space in the operating room.

  6. Flat panel detector-based cone beam computed tomography with a circle-plus-two-arcs data acquisition orbit: Preliminary phantom study

    International Nuclear Information System (INIS)

    Ning Ruola; Tang Xiangyang; Conover, David; Yu Rongfeng

    2003-01-01

    Cone beam computed tomography (CBCT) has been investigated in the past two decades due to its potential advantages over a fan beam CT. These advantages include (a) great improvement in data acquisition efficiency, spatial resolution, and spatial resolution uniformity, (b) substantially better utilization of x-ray photons generated by the x-ray tube compared to a fan beam CT, and (c) significant advancement in clinical three-dimensional (3D) CT applications. However, most studies of CBCT in the past are focused on cone beam data acquisition theories and reconstruction algorithms. The recent development of x-ray flat panel detectors (FPD) has made CBCT imaging feasible and practical. This paper reports a newly built flat panel detector-based CBCT prototype scanner and presents the results of the preliminary evaluation of the prototype through a phantom study. The prototype consisted of an x-ray tube, a flat panel detector, a GE 8800 CT gantry, a patient table and a computer system. The prototype was constructed by modifying a GE 8800 CT gantry such that both a single-circle cone beam acquisition orbit and a circle-plus-two-arcs orbit can be achieved. With a circle-plus-two-arcs orbit, a complete set of cone beam projection data can be obtained, consisting of a set of circle projections and a set of arc projections. Using the prototype scanner, the set of circle projections were acquired by rotating the x-ray tube and the FPD together on the gantry, and the set of arc projections were obtained by tilting the gantry while the x-ray tube and detector were at the 12 and 6 o'clock positions, respectively. A filtered backprojection exact cone beam reconstruction algorithm based on a circle-plus-two-arcs orbit was used for cone beam reconstruction from both the circle and arc projections. The system was first characterized in terms of the linearity and dynamic range of the detector. Then the uniformity, spatial resolution and low contrast resolution were assessed using

  7. Technical Note: Modification of the standard gain correction algorithm to compensate for the number of used reference flat frames in detector performance studies

    International Nuclear Information System (INIS)

    Konstantinidis, Anastasios C.; Olivo, Alessandro; Speller, Robert D.

    2011-01-01

    Purpose: The x-ray performance evaluation of digital x-ray detectors is based on the calculation of the modulation transfer function (MTF), the noise power spectrum (NPS), and the resultant detective quantum efficiency (DQE). The flat images used for the extraction of the NPS should not contain any fixed pattern noise (FPN) to avoid contamination from nonstochastic processes. The ''gold standard'' method used for the reduction of the FPN (i.e., the different gain between pixels) in linear x-ray detectors is based on normalization with an average reference flat-field. However, the noise in the corrected image depends on the number of flat frames used for the average flat image. The aim of this study is to modify the standard gain correction algorithm to make it independent on the used reference flat frames. Methods: Many publications suggest the use of 10-16 reference flat frames, while other studies use higher numbers (e.g., 48 frames) to reduce the propagated noise from the average flat image. This study quantifies experimentally the effect of the number of used reference flat frames on the NPS and DQE values and appropriately modifies the gain correction algorithm to compensate for this effect. Results: It is shown that using the suggested gain correction algorithm a minimum number of reference flat frames (i.e., down to one frame) can be used to eliminate the FPN from the raw flat image. This saves computer memory and time during the x-ray performance evaluation. Conclusions: The authors show that the method presented in the study (a) leads to the maximum DQE value that one would have by using the conventional method and very large number of frames and (b) has been compared to an independent gain correction method based on the subtraction of flat-field images, leading to identical DQE values. They believe this provides robust validation of the proposed method.

  8. Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT

    International Nuclear Information System (INIS)

    Struffert, Tobias; Kloska, Stephan; Engelhorn, Tobias; Doerfler, Arnd; Deuerling-Zheng, Yu; Boese, Jan; Zellerhoff, Michael; Schwab, Stefan

    2011-01-01

    We tested the hypothesis that Flat Detector computed tomography (FD-CT) with intravenous contrast medium would allow the calculation of whole brain cerebral blood volume (CBV) mapping (FD-CBV) and would correlate with multislice Perfusion CT (PCT). Twenty five patients were investigated with FD-CBV and PCT. Correlation of the CBV maps of both techniques was carried out with measurements from six anatomical regions from both sides of the brain. Mean values of each region and the correlation coefficient were calculated. Bland-Altman analysis was performed to compare the two different imaging techniques. The image and data quality of both PCT and FD-CBV were suitable for evaluation in all patients. The mean CBV values of FD-CBV and PCT showed only minimal differences with overlapping standard deviation. The correlation coefficient was 0.79 (p < 0.01). Bland-Altman analysis showed a mean difference of -0.077 ± 0.48 ml/100 g between FD-CBV and PCT CBV measurements, indicating that FD-CBV values were only slightly lower than those of PCT. CBV mapping with intravenous contrast medium using Flat Detector CT compared favourably with multislice PCT. The ability to assess cerebral perfusion within the angiographic suite may improve the management of ischaemic stroke and evaluation of the efficacy of dedicated therapies. (orig.)

  9. Robot arm based flat panel CT-guided electromagnetic tracked spine interventions: phantom and animal model experiments.

    Science.gov (United States)

    Penzkofer, Tobias; Isfort, Peter; Bruners, Philipp; Wiemann, Christian; Kyriakou, Yiannis; Kalender, Willi A; Günther, Rolf W; Schmitz-Rode, Thomas; Mahnken, Andreas H

    2010-11-01

    To evaluate accuracy and procedure times of electromagnetic tracking (EMT) in a robotic arm mounted flat panel setting using phantom and animal cadaveric models. A robotic arm mounted flat panel (RMFP) was used in combination with EMT to perform anthropomorphic phantom (n = 90) and ex vivo pig based punctures (n = 120) of lumbar facet joints (FJ, n = 120) and intervertebral discs (IVD, n = 90). Procedure accuracies and times were assessed and evaluated. FJ punctures were carried out with a spatial accuracy of 0.8 ± 0.9 mm (phantom) and 0.6 ± 0.8 mm (ex vivo) respectively. While IVD punctures showed puncture deviations of 0.6 ± 1.2 mm (phantom) and 0.5 ± 0.6 mm (ex vivo), direct and angulated phantom based punctures had accuracies of 0.8 ± 0.9 mm and 1.0 ± 1.3 mm. Planning took longer for ex vivo IVD punctures compared to phantom model interventions (39.3 ± 17.3 s vs. 20.8 ± 5.0 s, p = 0.001) and for angulated vs. direct phantom FJ punctures (19.7 ± 5.1 s vs. 28.6 ± 7.8 s, p < 0.001). Puncture times were longer for ex vivo procedures when compared to phantom model procedures in both FJ (37.9 ± 9.0 s vs. 23.6 ± 7.2 s, p = 0.001) and IVD punctures (43.9 ± 16.1 s vs. 31.1 ± 6.4 s, p = 0.026). The combination of RMFP with EMT provides an accurate method of navigation for spinal interventions such as facet joint punctures and intervertebral disc punctures.

  10. TU-AB-204-01: Advances in C-Arm CBCT for Brain Perfusion Imaging

    International Nuclear Information System (INIS)

    Chen, G.

    2015-01-01

    , significant effort has been expended to improve the quantitative accuracy of C-arm CBCT reconstructions. The challenge is to improve image quality while providing very short turnaround between data acquisition and volume data visualization. Corrections for x-ray scatter, view aliasing and patient motion that require no more than 2 iterations keep processing time short while reducing artifact. Fast, multi-sweep acquisitions can be used to permit assessment of left ventricular function, and visualization of radiofrequency lesions created to treat arrhythmias. Workflows for each imaging goal have been developed and validated against gold standard clinical CT or histology. The challenges, opportunities, and limitations of the new functional C-arm CBCT imaging techniques will be discussed. Dr. W. Zbijewski (Johns Hopkins University) will present on the topic: Advances in CBCT for Orthopaedics and Bone Health Imaging. Cone-beam CT is particularly well suited for imaging of musculoskeletal extremities. Owing to the high spatial resolution of flat-panel detectors, CBCT can surpass conventional CT in imaging tasks involving bone visualization, quantitative analysis of subchondral trabecular structure, and visualization and monitoring of subtle fractures that are common in orthopedic radiology. A dedicated CBCT platform has been developed that offers flexibility in system design and provides not only a compact configuration with improved logistics for extremities imaging but also enables novel diagnostic capabilities such as imaging of weight-bearing lower extremities in a natural stance. The design, development and clinical performance of dedicated extremities CBCT systems will be presented. Advanced capabilities for quantitative volumetric assessment of joint space morphology, dual-energy image-based quantification of bone composition, and in-vivo analysis of bone microarchitecture will be discussed, along with emerging applications in the diagnosis of arthritis and osteoporosis and

  11. TU-AB-204-01: Advances in C-Arm CBCT for Brain Perfusion Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Chen, G. [University of Wisconsin (United States)

    2015-06-15

    , significant effort has been expended to improve the quantitative accuracy of C-arm CBCT reconstructions. The challenge is to improve image quality while providing very short turnaround between data acquisition and volume data visualization. Corrections for x-ray scatter, view aliasing and patient motion that require no more than 2 iterations keep processing time short while reducing artifact. Fast, multi-sweep acquisitions can be used to permit assessment of left ventricular function, and visualization of radiofrequency lesions created to treat arrhythmias. Workflows for each imaging goal have been developed and validated against gold standard clinical CT or histology. The challenges, opportunities, and limitations of the new functional C-arm CBCT imaging techniques will be discussed. Dr. W. Zbijewski (Johns Hopkins University) will present on the topic: Advances in CBCT for Orthopaedics and Bone Health Imaging. Cone-beam CT is particularly well suited for imaging of musculoskeletal extremities. Owing to the high spatial resolution of flat-panel detectors, CBCT can surpass conventional CT in imaging tasks involving bone visualization, quantitative analysis of subchondral trabecular structure, and visualization and monitoring of subtle fractures that are common in orthopedic radiology. A dedicated CBCT platform has been developed that offers flexibility in system design and provides not only a compact configuration with improved logistics for extremities imaging but also enables novel diagnostic capabilities such as imaging of weight-bearing lower extremities in a natural stance. The design, development and clinical performance of dedicated extremities CBCT systems will be presented. Advanced capabilities for quantitative volumetric assessment of joint space morphology, dual-energy image-based quantification of bone composition, and in-vivo analysis of bone microarchitecture will be discussed, along with emerging applications in the diagnosis of arthritis and osteoporosis and

  12. Aging of imaging properties of a CMOS flat-panel detector for dental cone-beam computed tomography

    Science.gov (United States)

    Kim, D. W.; Han, J. C.; Yun, S.; Kim, H. K.

    2017-01-01

    We have experimentally investigated the long-term stability of imaging properties of a flat-panel detector in conditions used for dental x-ray imaging. The detector consists of a CsI:Tl layer and CMOS photodiode pixel arrays. Aging simulations were carried out using an 80-kVp x-ray beam at an air-kerma rate of approximately 5 mGy s-1 at the entrance surface of the detector with a total air kerma of up to 0.6 kGy. Dark and flood-field images were periodically obtained during irradiation, and the mean signal and noise levels were evaluated for each image. We also evaluated the modulation-transfer function (MTF), noise-power spectrum (NPS), and detective quantum efficiency (DQE). The aging simulation showed a decrease in both the signal and noise of the gain-offset-corrected images, but there was negligible change in the signal-to-noise performance as a function of the accumulated dose. The gain-offset correction for analyzing images resulted in negligible changes in MTF, NPS, and DQE results over the total dose. Continuous x-ray exposure to a detector can cause degradation in the physical performance factors such the detector sensitivity, but linear analysis of the gain-offset-corrected images can assure integrity of the imaging properties of a detector during its lifetime.

  13. The LiC detector toy program

    International Nuclear Information System (INIS)

    Regler, Meinhard; Valentan, Manfred; Fruehwirth, Rudolf

    2007-01-01

    This note describes the 'LiC Detector Toy' ('LiC' for Linear Collider) software tool which has been developed for detector design studies, aiming at investigating the resolution of reconstructed track parameters for the purpose of comparing and optimizing various detector setups. It consists of a simplified simulation of the detector measurements, taking into account multiple scattering, followed by full single track reconstruction using the Kalman filter. The tool is written in MATLAB and may be installed on a laptop. It can easily be used as a black-box tool by non-experts, but also adapted to individual needs

  14. Examination of C/C flat tile mock-ups with hypervapotron cooling after high heat flux testing

    International Nuclear Information System (INIS)

    Schedler, B.; Friedrich, T.; Traxler, H.; Eidenberger, E.; Scheu, C.; Clemens, H.; Pippan, R.; Escourbiac, F.

    2007-01-01

    Two C/C flat tile mock-ups with a hypervapotron cooling concept, have been successfully tested beyond ITER specification (3000 cycles at 15 MW/m 2 , 300 cycles at 20 MW/m 2 and 800-1000 cycles at 25 MW/m 2 ) in two electron beam testing facilities [F. Escourbiac, et al., Experimental simulation of cascade failure effect on tungsten and CFC flat tile armoured HHF components, Fusion Eng. Des., submitted for publication; F. Escourbiac, et al., A mature industrial solution for ITER divertor plasma facing components: hypervapotron cooling concept adapted to Tore Supra flat tile technology, Fusion Eng. Des. 75-79 (2005) 387-390]. Both mock-ups provide a SNECMA SEPCARB NS31 armour, which has been joined onto the CuCrZr heat sink by active metal casting (AMC) and electron beam welding (EBW). No tile detachment or sudden loss of single tiles has been observed; a cascade-like failure of flat tile armours was impossible to generate. At the maximum cyclic heat flux load of 25 MW/m 2 all tested tiles performed well except one, which revealed already a clear indication in the thermographic examination at the end of the manufacture. Visual examination and analysis of metallographic cuts of the remaining tiles demonstrated that the interface has not been altered. In addition, the shear strength of the C/C to copper joints measured after the high heat flux (HHF) test has been found to be still above the interlamellar shear strength of the used C/C material. The high resistance of the interface is explained by a modification of the C/C to copper joint interface due to silicon originating from the used C/C material

  15. Examination of C/C flat tile mock-ups with hypervapotron cooling after high heat flux testing

    Energy Technology Data Exchange (ETDEWEB)

    Schedler, B. [Technology Centre of PLANSEE SE, A-6600 Reutte (Austria)], E-mail: bertram.schedler@plansee.com; Friedrich, T.; Traxler, H. [Technology Centre of PLANSEE SE, A-6600 Reutte (Austria); Eidenberger, E.; Scheu, C.; Clemens, H. [Department of Physical Metallurgy and Materials Testing, University of Leoben, A-8700 Leoben (Austria); Pippan, R. [Austrian Academy of Sciences, Erich-Schmid-Institute of Material Science, A-8700 Leoben (Austria); Escourbiac, F. [Association EURATOM-CEA, DSM/DRFC, CEA Cadarache, F-13108 St. Paul Lez Durance (France)

    2007-04-15

    Two C/C flat tile mock-ups with a hypervapotron cooling concept, have been successfully tested beyond ITER specification (3000 cycles at 15 MW/m{sup 2}, 300 cycles at 20 MW/m{sup 2} and 800-1000 cycles at 25 MW/m{sup 2}) in two electron beam testing facilities [F. Escourbiac, et al., Experimental simulation of cascade failure effect on tungsten and CFC flat tile armoured HHF components, Fusion Eng. Des., submitted for publication; F. Escourbiac, et al., A mature industrial solution for ITER divertor plasma facing components: hypervapotron cooling concept adapted to Tore Supra flat tile technology, Fusion Eng. Des. 75-79 (2005) 387-390]. Both mock-ups provide a SNECMA SEPCARB NS31 armour, which has been joined onto the CuCrZr heat sink by active metal casting (AMC) and electron beam welding (EBW). No tile detachment or sudden loss of single tiles has been observed; a cascade-like failure of flat tile armours was impossible to generate. At the maximum cyclic heat flux load of 25 MW/m{sup 2} all tested tiles performed well except one, which revealed already a clear indication in the thermographic examination at the end of the manufacture. Visual examination and analysis of metallographic cuts of the remaining tiles demonstrated that the interface has not been altered. In addition, the shear strength of the C/C to copper joints measured after the high heat flux (HHF) test has been found to be still above the interlamellar shear strength of the used C/C material. The high resistance of the interface is explained by a modification of the C/C to copper joint interface due to silicon originating from the used C/C material.

  16. ACS/WFC Sky Flats from Frontier Fields Imaging

    Science.gov (United States)

    Mack, J.; Lucas, R. A.; Grogin, N. A.; Bohlin, R. C.; Koekemoer, A. M.

    2018-04-01

    Parallel imaging data from the HST Frontier Fields campaign (Lotz et al. 2017) have been used to compute sky flats for the ACS/WFC detector in order to verify the accuracy of the current set of flat field reference files. By masking sources and then co-adding many deep frames, the F606W and F814W filters have enough combined background signal that from Poisson statistics are efficiency tracks the thickness of the two WFC chips. Observations of blue and red calibration standards measured at various positions on the detector (Bohlin et al. 2017) confirm the fidelity of the F814W flat, with aperture photometry consistent to 1% across the FOV, regardless of spectral type. At bluer wavelengths, the total sky background is substantially lower, and the F435W sky flat shows a combination of both flat errors and detector artifacts. Aperture photometry of the red standard star shows a maximum deviation of 1.4% across the array in this filter. Larger residuals up to 2.5% are found for the blue standard, suggesting that the spatial sensitivity in F435W depends on spectral type.

  17. C-arm CT-guided renal arterial embolisation followed by radiofrequency ablation for treatment of patients with unresectable renal cell carcinoma

    International Nuclear Information System (INIS)

    Duan, X.-H.; Li, Y.-S.; Han, X.-W.; Wang, Y.-L.; Jiao, D.-C.; Li, T.-F.; Chen, P.-F.; Fang, Y.

    2016-01-01

    Aim: To explore the value of using flat detector (FD) equipped angiographic C-arm CT (CACT) systems in treating unresectable renal cell carcinoma (RCC) by selective renal arterial embolisation (RAE) followed by radiofrequency ablation (RFA) (RAE-RFA). Materials and methods: A total of 28 patients who were not candidates for surgery were enrolled. The average size of tumours was 6.7±2.2 cm (range 4.1–9.6 cm). Twenty-eight tumours were treated with CACT-guided RFA, 5–7 days after CACT-guided RAE. Results: CACT-guided RAE-RFA was technically successful in all patients. Tumour enhancement disappeared after a single RAE-RFA session in 20 patients, after two RAE-RFA sessions in four patients and after three RAE-RFA sessions in the other four patients. One patient died of lung metastasis and haematuria 13 months after RAE-RFA, and another patient died of pulmonary heart disease 23 months after repeat RAE-RFA. In the 26 living patients, tumours remained controlled during a mean follow-up period of 27 months and showed significant reduction in tumour size (6.7±2.2 cm to 3.9±1.7 cm, p<0.01). There were no significant changes in creatinine levels or urea nitrogen concentrations before and after the last RAE-RFA (p>0.05). There were no serious complications during and after the procedure. Conclusion: CACT-guided RAE followed by RFA appears to be a safe and effective technique for treating patients with inoperable RCC. - Highlights: • Selective RAE combined with RFA (RAE-RFA) is used to treat unresectable RCC. • We assessed the C-arm CT (CACT) based needle guidance system to perform RAE-RFA. • Tumors were controlled and reduced in size following CACT-guided RAE-RFA. • No significant changes in renal function occurred post-CACT-guided RAE-RFA. • CACT-guided RAE-RFA for treating RCC appears to be a safe and effective technique.

  18. Variation in X-ray dose quantity using an amorphous selenium based flat-panel detector - a study on the dose reduction rate up to the limit of diagnostical utilization

    International Nuclear Information System (INIS)

    Lehnert, T.; Wohlers, J.; Manegold, K.; Wetter, A.; Jacobi, V.; Mack, M.G.; Vogl, T.J.; Streng, W.

    2006-01-01

    Purpose: To evaluate the diagnostic quality and minimum required dose to obtain acceptable images for diagnostic purposes in the field of musculoskeletal radiology. Materials and methods: A critical comparison of the image quality produced by a novel flat panel detector and the conventional screen/film system using a contrast-detail phantom was performed in phase I. Images from both systems were obtained with the same dose and displayed with similar contrast and density. In phase II images of significant anatomical structures in cadaver extremities obtained using the digital detector system and the standard film/screen system were critically evaluated. After a successive reduction in the X-ray dose for 84 patients in phase III, eight independent radiologists compared the image quality of the screen/film system to that of the novel flat panel detector. Results: Phases I and II revealed a difference in the image quality achieved by the standard screen/film system and the digital detector system to the advantage of the digital detector system. In 77 of 84 patients (91.7%), phase III showed equal image quality after a 50% reduction in the X-ray dose. In 3 cases (3.6%) the image quality and the level of contrast were better. No unified statement could be made for 4 patients (4.7%). Conclusion: Digital imaging of skeletal disorders using the novel flat panel detector makes it possible to reduce the X-ray dose by 50% with equal or even better image quality. (orig.)

  19. Quantitative comparison using Generalized Relative Object Detectability (G-ROD) metrics of an amorphous selenium detector with high resolution Microangiographic Fluoroscopes (MAF) and standard flat panel detectors (FPD).

    Science.gov (United States)

    Russ, M; Shankar, A; Jain, A; Setlur Nagesh, S V; Ionita, C N; Scott, C; Karim, K S; Bednarek, D R; Rudin, S

    2016-02-27

    A novel amorphous selenium (a-Se) direct detector with CMOS readout has been designed, and relative detector performance investigated. The detector features include a 25 μ m pixel pitch, and 1000 μ m thick a-Se layer operating at 10V/ μ m bias field. A simulated detector DQE was determined, and used in comparative calculations of the Relative Object Detectability (ROD) family of prewhitening matched-filter (PWMF) observer and non-prewhitening matched filter (NPWMF) observer model metrics to gauge a-Se detector performance against existing high resolution micro-angiographic fluoroscopic (MAF) detectors and a standard flat panel detector (FPD). The PWMF-ROD or ROD metric compares two x-ray imaging detectors in their relative abilities in imaging a given object by taking the integral over spatial frequencies of the Fourier transform of the detector DQE weighted by an object function, divided by the comparable integral for a different detector. The generalized-ROD (G-ROD) metric incorporates clinically relevant parameters (focal-spot size, magnification, and scatter) to show the degradation in imaging performance for detectors that are part of an imaging chain. Preliminary ROD calculations using simulated spheres as the object predicted superior imaging performance by the a-Se detector as compared to existing detectors. New PWMF-G-ROD and NPWMF-G-ROD results still indicate better performance by the a-Se detector in an imaging chain over all sphere sizes for various focal spot sizes and magnifications, although a-Se performance advantages were degraded by focal spot blurring. Nevertheless, the a-Se technology has great potential to provide breakthrough abilities such as visualization of fine details including of neuro-vascular perforator vessels and of small vascular devices.

  20. Performance Study of an aSi Flat Panel Detector for Fast Neutron Imaging of Nuclear Waste

    Energy Technology Data Exchange (ETDEWEB)

    Schumann, M.; Mauerhofer, E. [Institute of Energy and Climate Research - Nuclear Waste Management and Reactor Safety, Forschungszentrum Juelich GmbH, 52425 Juelich (Germany); Engels, R.; Kemmerling, G. [Central Institute for Engineering, Electronics and Analytics - Electronic Systems, Forschungszentrum Juelich GmbH, 52425 Juelich (Germany); Frank, M. [MATHCCES - Department of Mathematics, RWTH Aachen University, 52062 Aachen (Germany); Havenith, A.; Kettler, J.; Klapdor-Kleingrothaus, T. [Institute of Nuclear Engineering and Technology Transfer, RWTH Aachen University, 52062 Aachen (Germany); Schitthelm, O. [Corporate Technology, Siemens AG, 91058 Erlangen (Germany)

    2015-07-01

    Radioactive waste must be characterized to check its conformance for intermediate storage and final disposal according to national regulations. For the determination of radio-toxic and chemo-toxic contents of radioactive waste packages non-destructive analytical techniques are preferentially used. Fast neutron imaging is a promising technique to assay large and dense items providing, in complementarity to photon imaging, additional information on the presence of structures in radioactive waste packages. Therefore the feasibility of a compact Neutron Imaging System for Radioactive waste Analysis (NISRA) using 14 MeV neutrons is studied in a cooperation framework of Forschungszentrum Juelich GmbH, RWTH Aachen University and Siemens AG. However due to the low neutron emission of neutron generators in comparison to research reactors the challenging task resides in the development of an imaging detector with a high efficiency, a low sensitivity to gamma radiation and a resolution sufficient for the purpose. The setup is composed of a commercial D-T neutron generator (Genie16GT, Sodern) with a surrounding shielding made of polyethylene, which acts as a collimator and an amorphous silicon flat panel detector (aSi, 40 x 40 cm{sup 2}, XRD-1642, Perkin Elmer). Neutron detection is achieved using a general propose plastic scintillator (EJ-260, Eljen Technology) linked to the detector. The thermal noise of the photodiodes is reduced by employing an entrance window made of aluminium. Optimal gain and integration time for data acquisition are set by measuring the response of the detector to the radiation of a 500 MBq {sup 241}Am-source. Detector performance was studied by recording neutron radiography images of materials with various, but well known, chemical compositions, densities and dimensions (Al, C, Fe, Pb, W, concrete, polyethylene, 5 x 8 x 10 cm{sup 3}). To simulate gamma-ray emitting waste radiographs in presence of a gamma-ray sources ({sup 60}Co, {sup 137}Cs, {sup 241

  1. Comparison of dose and image quality of a Flat-panel detector and an image intensifier; Comparacao da dose e qualidade da imagem de um detector Flatpanel e um intensificador de imagem

    Energy Technology Data Exchange (ETDEWEB)

    Lazzaro, M.; Friedrich, B.Q.; Luz, R.M. da; Silva, A.M.M. da, E-mail: marcos.lazzaro@acad.pucrs.br [Pontificia Universidade Catolica do Rio Grande do Sul (PUC-RS), Porto Alegre, RS (Brazil)

    2016-07-01

    With the development of new technologies, have emerged new conversion methods of X-ray image, such as flat panel detectors. The aim of this work is the comparison of entrance surface air kerma (ESAK) and image quality between an image intensifier type of detector (A) and a flat panel (B). The ESAK was obtained by placing a ionization chamber under PMMA simulators of 10, 20 and 30 cm and the image quality was obtained by using the TOR {sup 18}FG simulator. The ESAK to the equipment A is higher when compared to the equipment B. The high contrast resolution is better for the equipment A for all thicknesses of simulators. The equipment A has low contrast resolution with a better viewing threshold for thicknesses of 10 and 20 cm, and a worse performance for 30 cm. It is concluded that the equipment B has ESAK smaller and despite having lower resolution, in almost all cases, have appropriate image quality for diagnosis. (author)

  2. Unequal-Arms Michelson Interferometers

    Science.gov (United States)

    Tinto, Massimo; Armstrong, J. W.

    2000-01-01

    Michelson interferometers allow phase measurements many orders of magnitude below the phase stability of the laser light injected into their two almost equal-length arms. If, however, the two arms are unequal, the laser fluctuations can not be removed by simply recombining the two beams. This is because the laser jitters experience different time delays in the two arms, and therefore can not cancel at the photo detector. We present here a method for achieving exact laser noise cancellation, even in an unequal-arm interferometer. The method presented in this paper requires a separate readout of the relative phase in each arm, made by interfering the returning beam in each arm with a fraction of the outgoing beam. By linearly combining the two data sets with themselves, after they have been properly time shifted, we show that it is possible to construct a new data set that is free of laser fluctuations. An application of this technique to future planned space-based laser interferometer detector3 of gravitational radiation is discussed.

  3. Extra-oral dental radiography for disaster victims using a flat panel X-ray detector and a hand-held X-ray generator.

    Science.gov (United States)

    Ohtani, M; Oshima, T; Mimasaka, S

    2017-12-01

    Forensic odontologists commonly incise the skin for post-mortem dental examinations when it is difficult to open the victim's mouth. However, it is prohibited by law to incise dead bodies without permission in Japan. Therefore, we attempted using extra-oral dental radiography, using a digital X-ray equipment with rechargeable batteries, to overcome this restriction. A phantom was placed in the prone position on a table, and three plain dental radiographs were used per case: "lateral oblique radiographs" for left and right posterior teeth and a "contact radiograph" for anterior teeth were taken using a flat panel X-ray detector and a hand-held X-ray generator. The resolving power of the images was measured by a resolution test chart, and the scattered X-ray dose was measured using an ionization chamber-type survey meter. The resolving power of the flat panel X-ray detector was 3.0 lp/mm, which was less than that of intra-oral dental methods, but the three extra-oral plain dental radiographs provided the overall dental information from outside of the mouth, and this approach was less time-consuming. In addition, the higher dose of scattered X-rays was laterally distributed, but the dose per case was much less than that of intra-oral dental radiographs. Extra-oral plain dental radiography can be used for disaster victim identification by dental methods even when it is difficult to open the mouth. Portable and rechargeable devices, such as a flat panel X-ray detector and a hand-held X-ray generator, are convenient to bring and use anywhere, even at a disaster scene lacking electricity and water.

  4. C-arm guided closed reduction of zygomatic arch fracture

    International Nuclear Information System (INIS)

    Eo, Yoon Ki; Lee, Dong Kun; Kim, Jeong Sam; Jang, Young Il

    1999-01-01

    The zygomatic arch is structurally protruded and is easily fractured. The classic management of zygomatic arch fracture has been mentioned the Keen, Lothrop, Dingman and Alling and threaded K-wire. All of the above methods have advantages and disadvantages. To minimize the disadvantages, we performed threaded K-wire for the first time using C-arm image intensifier. The subjects were 16 patients with Knight North group II (Zygomatic arch fracture). Among them the C-arm was used in 12 patients and the operator used sensitivity general method in 4 patients and confirmed the operation by mobile X-ray equipment. In conclusion, both groups were satisfied surgically and cosmetically. Using the C-arm, actual image at the time operation was clear and satisfied, the surrounding tissue damage was minimized and at was more accurately completed. The operation time was shortened by 30 to 60 minutes proving it to be an efficient method. We suggest though that further studies be needed to evaluate the radiation effect on these patients

  5. Remarks on the conformally flat submanifolds satisfying the R.C-condition or the C.R-condition

    International Nuclear Information System (INIS)

    Zafindratafa, G.K.

    1988-06-01

    The hypersurfaces of E n+1 have been studied for the particular case when they satisfy the R.C-condition or the C.R-condition. One objective is to generalize this situation to a higher codimension. More precisely, we consider the case of dimension 4, and replace the condition of quasiumbilicity by the conformal flatness. In this way, we construct an example of 4-submanifold of IE 6 which is conformally flat at a particular point without being quasiumbilical. That such submanifolds exist, was asserted without proof. Thus, we present another counter-example. (author). 4 refs

  6. Measurements of surgeons' exposure to ionizing radiation dose: comparison of conventional and mini C-arm fluoroscopy.

    Science.gov (United States)

    Sung, K H; Min, E; Chung, C Y; Jo, B C; Park, M S; Lee, K

    2016-03-01

    This study was performed to measure the equivalent scattered radiation dose delivered to susceptible organs while simulating orthopaedic surgery using conventional and mini C-arm fluoroscopy. In addition, shielding effects on the thyroid, thymus, and gonad, and the direct exposure delivered to the patient's hands were also compared. A conventional and mini C-arms were installed in an operating room, and a hand and an operator phantom were used to simulate a patient's hand and a surgeon. Photoluminescence dosimeters were used to measure the equivalent dose by scattered radiation arriving at the thyroid, thymus, and gonad on a whole-body phantom in the position of the surgeon. Equivalent scattered radiation doses were measured in four groups: (1) unshielded conventional C-arm group; (2) unshielded mini C-arm group; (3) lead-shielded conventional C-arm group; and (4) lead-shielded mini C-arm group. Equivalent scattered radiation doses to the unshielded group were significantly lower in the mini C-arm group than those in the conventional C-arm group for all organs. The gonad in the lead-shielded conventional C-arm group showed the highest equivalent dose among operator-susceptible organs, and radiation dose was reduced by approximately 96% compared with that in the unshielded group. Scattered radiation was not detected in any susceptible organ in the lead-shielded mini C-arm group. The direct radiation dose to the hand phantom measured from the mini C-arm was significantly lower than that measured from the conventional C-arm. The results show that the equivalent scattered radiation dose to the surgeon's susceptible organs and the direct radiation dose to a patient's hand can be decreased significantly by using a mini C-arm rather than a conventional C-arm. However, protective lead garments, such as a thyroid shield and apron, should be applied to minimize radiation exposure to susceptible organs, even during use of mini C-arm fluoroscopy. © The Author(s) 2015.

  7. A forward bias method for lag correction of an a-Si flat panel detector

    International Nuclear Information System (INIS)

    Starman, Jared; Tognina, Carlo; Partain, Larry; Fahrig, Rebecca

    2012-01-01

    Purpose: Digital a-Si flat panel (FP) x-ray detectors can exhibit detector lag, or residual signal, of several percent that can cause ghosting in projection images or severe shading artifacts, known as the radar artifact, in cone-beam computed tomography (CBCT) reconstructions. A major contributor to detector lag is believed to be defect states, or traps, in the a-Si layer of the FP. Software methods to characterize and correct for the detector lag exist, but they may make assumptions such as system linearity and time invariance, which may not be true. The purpose of this work is to investigate a new hardware based method to reduce lag in an a-Si FP and to evaluate its effectiveness at removing shading artifacts in CBCT reconstructions. The feasibility of a novel, partially hardware based solution is also examined. Methods: The proposed hardware solution for lag reduction requires only a minor change to the FP. For pulsed irradiation, the proposed method inserts a new operation step between the readout and data collection stages. During this new stage the photodiode is operated in a forward bias mode, which fills the defect states with charge. A Varian 4030CB panel was modified to allow for operation in the forward bias mode. The contrast of residual lag ghosts was measured for lag frames 2 and 100 after irradiation ceased for standard and forward bias modes. Detector step response, lag, SNR, modulation transfer function (MTF), and detective quantum efficiency (DQE) measurements were made with standard and forward bias firmware. CBCT data of pelvic and head phantoms were also collected. Results: Overall, the 2nd and 100th detector lag frame residual signals were reduced 70%-88% using the new method. SNR, MTF, and DQE measurements show a small decrease in collected signal and a small increase in noise. The forward bias hardware successfully reduced the radar artifact in the CBCT reconstruction of the pelvic and head phantoms by 48%-81%. Conclusions: Overall, the

  8. Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: Visibility of simulated microcalcifications

    OpenAIRE

    Shen, Youtao; Zhong, Yuncheng; Lai, Chao-Jen; Wang, Tianpeng; Shaw, Chris C.

    2013-01-01

    Purpose: To measure and investigate the improvement of microcalcification (MC) visibility in cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS/CsI flat panel detector (Dexela 2923, Perkin Elmer).

  9. LUCID A Cherenkov Tube Based Detector for Monitoring the ATLAS Experiment Luminosity

    CERN Document Server

    Sbrizzi, A

    2007-01-01

    The LUCID (LUminosity Cherenkov Integrating Detector) apparatus is composed by two symmetric arms deployed at about 17 m from the ATLAS interaction point. The purpose of this detector, which will be installed in january 2008, is to monitor the luminosity delivered by the LHC machine to the ATLAS experiment. An absolute luminosity calibration is needed and it will be provided by a Roman Pot type detector with the two arms placed at about 240 m from the interaction point. Each arm of the LUCID detector is based on an aluminum vessel containing 20 Cherenkov tubes, 15 mm diameter and 1500 mm length, filled with C4F10 radiator gas at 1.5 bar. The Cherenkov light generated by charged particles above the threshold is collected by photomultiplier tubes (PMT) directly placed at the tubes end. The challenging aspect of this detector is its readout in an environment characterized by the high dose of radiation (about 0.7 Mrad/year at 10^33cm^2 s^-1) it must withstand. In order to fulfill these radiation hardness requirem...

  10. Source strength verification and quality assurance of preloaded brachytherapy needles using a CMOS flat panel detector

    Energy Technology Data Exchange (ETDEWEB)

    Golshan, Maryam, E-mail: maryam.golshan@bccancer.bc.ca [Department of Physics, University of British Columbia, Vancouver, British Columbia V6T1Z1, Canada and Department of Medical Physics, Vancouver Center, British Columbia Cancer Agency, Vancouver, British Columbia V5Z 4E6 (Canada); Spadinger, Ingrid [Department of Medical Physics, Vancouver Center, British Columbia Cancer Agency, Vancouver, British Columbia V5Z 4E6 (Canada); Chng, Nick [Department of Medical Physics, Center for the North, British Columbia Cancer Agency, Prince George, British Columbia V2M 7E9 (Canada)

    2016-06-15

    Purpose: Current methods of low dose rate brachytherapy source strength verification for sources preloaded into needles consist of either assaying a small number of seeds from a separate sample belonging to the same lot used to load the needles or performing batch assays of a subset of the preloaded seed trains. Both of these methods are cumbersome and have the limitations inherent to sampling. The purpose of this work was to investigate an alternative approach that uses an image-based, autoradiographic system capable of the rapid and complete assay of all sources without compromising sterility. Methods: The system consists of a flat panel image detector, an autoclavable needle holder, and software to analyze the detected signals. The needle holder was designed to maintain a fixed vertical spacing between the needles and the image detector, and to collimate the emissions from each seed. It also provides a sterile barrier between the needles and the imager. The image detector has a sufficiently large image capture area to allow several needles to be analyzed simultaneously.Several tests were performed to assess the accuracy and reproducibility of source strengths obtained using this system. Three different seed models (Oncura 6711 and 9011 {sup 125}I seeds, and IsoAid Advantage {sup 103}Pd seeds) were used in the evaluations. Seeds were loaded into trains with at least 1 cm spacing. Results: Using our system, it was possible to obtain linear calibration curves with coverage factor k = 1 prediction intervals of less than ±2% near the centre of their range for the three source models. The uncertainty budget calculated from a combination of type A and type B estimates of potential sources of error was somewhat larger, yielding (k = 1) combined uncertainties for individual seed readings of 6.2% for {sup 125}I 6711 seeds, 4.7% for {sup 125}I 9011 seeds, and 11.0% for Advantage {sup 103}Pd seeds. Conclusions: This study showed that a flat panel detector dosimetry system

  11. A comparison of entrance skin dose delivered by clinical angiographic c-arms using the real-time dosimeter: the MOSkin

    International Nuclear Information System (INIS)

    Thorpe, Nathan K.; Cutajar, Dean; Lian, Cheryl; Rosenfeld, Anatoly; Pitney, Mark; Friedman, Daniel; Perevertaylo, Vladimir

    2016-01-01

    Coronary angiography is a procedure used in the diagnosis and intervention of coronary heart disease. The procedure is often considered one of the highest dose diagnostic procedures in clinical use. Despite this, there is minimal use of dosimeters within angiographic catheterisation laboratories due to challenges resulting from their implementation. The aim of this study was to compare entrance dose delivery across locally commissioned c-arms to assess the need for real-time dosimetry solutions during angiographic procedures. The secondary aim of this study was to establish a calibration method for the MOSkin dosimeter that accurately produces entrance dose values from the clinically sampled beam qualities and energies. The MOSkin is a real-time dosimeter used to measure the skin dose delivered by external radiation beams. The suitability of the MOSkin for measurements in the angiographic catheterisation laboratory was assessed. Measurements were performed using a 30 × 30 × 30cm 3 PMMA phantom positioned at the rotational isocenter of the c-arm gantry. The MOSkin calibration factor was established through comparison of the MOSkin response to EBT2 film response. Irradiation of the dosimeters was performed using several clinical beam qualities ranging in energy from 70 to 105 kVp. A total of four different interventional c-arm machines were surveyed and compared using the MOSkin dosimeter. The phantom was irradiated from a normal angle of incidence using clinically relevant protocols, field sizes and source to image detector distance values. The MOSkin was observed to be radiotranslucent to the c-arm beam in all clinical environments. The MOSkin response was reproducible to within 2 % of the average value across repeated measurements for each beam setting. There were large variations in entrance dose delivery to the phantom between the different c-arm machines with the highest observed cine-acquisition entrance dose rate measuring 326 % higher than the lowest

  12. A new index of flat breaststroke propulsion: a comparison of elite men and women.

    Science.gov (United States)

    Seifert, L; Chollet, D

    2005-03-01

    This study examined arm and leg coordination and propulsion during the flat breaststroke in nine elite male and eight elite female swimmers over three race paces (200 m, 100 m and 50 m). Coordination was expressed using four temporal gaps (T1, T2, T3, T4), which described the continuity between the propulsive phases of the limbs, as recorded on a video device (50 Hz). Glide duration was denoted T1, the time between the beginning of arm and leg recovery was denoted T2, the time between the end of arm and the leg recovery was denoted T3, and the time between 90 degrees of flexion during arm recovery and 90 degrees during leg recovery was denoted T4. Using these temporal gaps, four stroke phases (propulsion, glide, recovery and leg insweep) could be followed over a complete arm and leg stroke. The total duration of arm and leg propulsion was assessed by a new index of flat breaststroke propulsion (IFBP). Velocity, stroke rate and stroke length were also calculated for each pace. The elite swimmers showed short T2, T3 and T4; moreover, T1 decreased when the pace increased. Expertise in the flat breaststroke was thus characterized by synchronized arm and leg recoveries and increased continuity in the arm and leg propulsions with increasing velocity. Differences between the sexes in the spatio-temporal parameters were possibly due to anthropometric differences (the men were heavier, older and taller than the women) and different motor organization linked to arm and leg coordination (shorter T3, body glide and body recovery, and greater body propulsion and higher IFBP in the men). The men's propulsive actions showed greater continuity, particularly in the sprint. The best men adopted a superposition coordination and thus had the ability to overcome very great active drag. Temporal gap measurement and the IFBP are practical indicators of arm and leg coordination and propulsion that can be exploited by coaches and swimmers to increase the continuity between propulsive

  13. Influence of a combined CT/C-arm system on periprocedural workflow and procedure times in mechanical thrombectomy.

    Science.gov (United States)

    Pfaff, Johannes; Schönenberger, Silvia; Herweh, Christian; Pham, Mirko; Nagel, Simon; Ringleb, Peter Arthur; Heiland, Sabine; Bendszus, Martin; Möhlenbruch, Markus Alfred

    2017-09-01

    To achieve the fastest possible workflow in ischaemic stroke, we developed a CT/C-arm system, which allows imaging and endovascular treatment on the same patient table. This prospective, monocentric trial was conducted between October 2014 and August 2016. Patients received stroke imaging and mechanical thrombectomy under general anaesthesia (GA) or conscious sedation (CS) using our combined setup comprising a CT-scanner and a mobile C-arm X-ray device. Primary endpoint was time between stroke imaging and groin puncture. We compared periprocedural workflow and procedure times with the literature and a matched patient cohort treated with a biplane angiographic system before installation of the CT/C-arm system. In 50 patients with acute ischaemic stroke due to large-vessel occlusion in the anterior circulation, comparable recanalization rates were achieved by using the CT/C-arm setup (TICI2b-3:CT/C-arm-GA: 85.7%; CT/C-arm-CS: 90.9%; Angiosuite: 78.6%; p = 0.269) without increasing periprocedural complications. Elimination of patient transport resulted in a significant reduction of the time between stroke imaging and groin puncture: median, min (IQR): CT/C-arm-GA: 43 (35-52); CT/C-arm-CS: 39 (28-49); Angiosuite: 64 (48-74); p < 0.0001. The combined CT/C-arm system allows comparable recanalization rates as a biplane angiographic system and accelerates the start of the endovascular stroke treatment. • The CT/C-arm setup reduces median time from stroke imaging to groin puncture. • Mechanical thrombectomy using a C-arm device is feasible without increasing peri-interventional complications. • The CT/C-arm setup might be a valuable fallback solution for emergency procedures. • The CT/C-arm setup allows immediate control CT images during and after treatment.

  14. The design and imaging characteristics of dynamic, solid-state, flat-panel x-ray image detectors for digital fluoroscopy and fluorography

    International Nuclear Information System (INIS)

    Cowen, A.R.; Davies, A.G.; Sivananthan, M.U.

    2008-01-01

    Dynamic, flat-panel, solid-state, x-ray image detectors for use in digital fluoroscopy and fluorography emerged at the turn of the millennium. This new generation of dynamic detectors utilize a thin layer of x-ray absorptive material superimposed upon an electronic active matrix array fabricated in a film of hydrogenated amorphous silicon (a-Si:H). Dynamic solid-state detectors come in two basic designs, the indirect-conversion (x-ray scintillator based) and the direct-conversion (x-ray photoconductor based). This review explains the underlying principles and enabling technologies associated with these detector designs, and evaluates their physical imaging characteristics, comparing their performance against the long established x-ray image intensifier television (TV) system. Solid-state detectors afford a number of physical imaging benefits compared with the latter. These include zero geometrical distortion and vignetting, immunity from blooming at exposure highlights and negligible contrast loss (due to internal scatter). They also exhibit a wider dynamic range and maintain higher spatial resolution when imaging over larger fields of view. The detective quantum efficiency of indirect-conversion, dynamic, solid-state detectors is superior to that of both x-ray image intensifier TV systems and direct-conversion detectors. Dynamic solid-state detectors are playing a burgeoning role in fluoroscopy-guided diagnosis and intervention, leading to the displacement of x-ray image intensifier TV-based systems. Future trends in dynamic, solid-state, digital fluoroscopy detectors are also briefly considered. These include the growth in associated three-dimensional (3D) visualization techniques and potential improvements in dynamic detector design

  15. The scintillating fiber focal plane detector for the use of Kaos as a double arm spectrometer

    Energy Technology Data Exchange (ETDEWEB)

    Ayerbe Gayoso, Carlos Antonio

    2012-05-25

    The upgrade of the Mainz Mikrotron (MAMI) electron accelerator facility in 2007 which raised the beam energy up to 1.5 GeV, gives the opportunity to study strangeness production channels through electromagnetic process. The Kaon Spectrometer (KAOS) managed by the A1 Collaboration, enables the efficient detection of the kaons associated with strangeness electroproduction. Used as a single arm spectrometer, it can be combined with the existing high-resolution spectrometers for exclusive measurements in the kinematic domain accessible to them. For studying hypernuclear production in the {sup A}Z(e,e'K{sup +}){sup A}{sub {lambda}}(Z-1) reaction, the detection of electrons at very forward angles is needed. Therefore, the use of KAOS as a double-arm spectrometer for detection of kaons and the electrons at the same time is mandatory. Thus, the electron arm should be provided with a new detector package, with high counting rate capability and high granularity for a good spatial resolution. To this end, a new state-of-the-art scintillating fiber hodoscope has been developed as an electron detector. The hodoscope is made of two planes with a total of 18432 scintillating double-clad fibers of 0.83 mm diameter. Each plane is formed by 72 modules. Each module is formed from a 60 slanted multi-layer bundle, where 4 fibers of a tilted column are connected to a common read out. The read-out is made with 32 channels of linear array multianode photomultipliers. Signal processing makes use of newly developed double-threshold discriminators. The discriminated signal is sent in parallel to dead-time free time-to-digital modules and to logic modules for triggering purposes. Two fiber modules were tested with a carbon beam at GSI, showing a time resolution of {proportional_to}220 ps (FWHM) and a position resolution of {proportional_to}270 {mu}m (FWHM) with a detection efficiency {epsilon}>99%. The characterization of the spectrometer arm has been achieved through simulations

  16. The scintillating fiber focal plane detector for the use of Kaos as a double arm spectrometer

    International Nuclear Information System (INIS)

    Ayerbe Gayoso, Carlos Antonio

    2012-01-01

    The upgrade of the Mainz Mikrotron (MAMI) electron accelerator facility in 2007 which raised the beam energy up to 1.5 GeV, gives the opportunity to study strangeness production channels through electromagnetic process. The Kaon Spectrometer (KAOS) managed by the A1 Collaboration, enables the efficient detection of the kaons associated with strangeness electroproduction. Used as a single arm spectrometer, it can be combined with the existing high-resolution spectrometers for exclusive measurements in the kinematic domain accessible to them. For studying hypernuclear production in the A Z(e,e'K + ) A Λ (Z-1) reaction, the detection of electrons at very forward angles is needed. Therefore, the use of KAOS as a double-arm spectrometer for detection of kaons and the electrons at the same time is mandatory. Thus, the electron arm should be provided with a new detector package, with high counting rate capability and high granularity for a good spatial resolution. To this end, a new state-of-the-art scintillating fiber hodoscope has been developed as an electron detector. The hodoscope is made of two planes with a total of 18432 scintillating double-clad fibers of 0.83 mm diameter. Each plane is formed by 72 modules. Each module is formed from a 60 slanted multi-layer bundle, where 4 fibers of a tilted column are connected to a common read out. The read-out is made with 32 channels of linear array multianode photomultipliers. Signal processing makes use of newly developed double-threshold discriminators. The discriminated signal is sent in parallel to dead-time free time-to-digital modules and to logic modules for triggering purposes. Two fiber modules were tested with a carbon beam at GSI, showing a time resolution of ∝220 ps (FWHM) and a position resolution of ∝270 μm (FWHM) with a detection efficiency ε>99%. The characterization of the spectrometer arm has been achieved through simulations calculating the transfer matrix of track parameters from the fiber

  17. Intravenous flat-detector CT angiography in acute ischemic stroke management

    Energy Technology Data Exchange (ETDEWEB)

    Blanc, Raphael; Pistocchi, Silvia; Bartolini, Bruno; Piotin, Michel [Fondation Rothschild Hospital, Department of Interventional Neuroradiology, Paris (France); Babic, Drazenko [Philips Healthcare, Best (Netherlands); Obadia, Michael [Fondation Rothschild Hospital, Department of Neurology, Paris (France); Alamowitch, Sonia [APHP Hopital Tenon, Universite Paris VI, Department of Neurology, Paris (France)

    2012-04-15

    In the settings of stroke, a non-invasive high-resolution imaging modality to visualize the arterial intracranial circulation in the interventional lab is a helpful mean to plan the endovascular recanalization procedure. We report our initial experience with intravenously enhanced flat-detector CT (IV FDCT) technology in the detection of obstructed intracranial arteries. Fourteen consecutive patients elected for endovascular stroke therapy underwent IV FDCT. The scans were intravenously enhanced and acquired in accordance with the previously calculated bolus arrival time. Images were processed on a commercially available workstation for reconstructions and 3D manipulation. Occlusion level and clot length, the quality of collateral vessels, and the patency of anterior and posterior communicating arteries were assessed. IV FDCT was performed successfully in all the cases and allowed for clot location and length visualization, assessment of communicating arteries patency, and evaluation of vessel collateral grade. Information obtained from this technique was considered useful for patients treated by endovascular approach. Retrospective review of the images by two independent readers was considered accurate and reproducible. IV FDCT technology provided accurate delineation of obstructed vessel segments in acute ischemic stroke disease. It gave a significant help in the interventional strategy. This new technology available in the operating room might provide a valuable tool in emerging endovascular stroke therapy. (orig.)

  18. Intravenous flat-detector CT angiography in acute ischemic stroke management

    International Nuclear Information System (INIS)

    Blanc, Raphael; Pistocchi, Silvia; Bartolini, Bruno; Piotin, Michel; Babic, Drazenko; Obadia, Michael; Alamowitch, Sonia

    2012-01-01

    In the settings of stroke, a non-invasive high-resolution imaging modality to visualize the arterial intracranial circulation in the interventional lab is a helpful mean to plan the endovascular recanalization procedure. We report our initial experience with intravenously enhanced flat-detector CT (IV FDCT) technology in the detection of obstructed intracranial arteries. Fourteen consecutive patients elected for endovascular stroke therapy underwent IV FDCT. The scans were intravenously enhanced and acquired in accordance with the previously calculated bolus arrival time. Images were processed on a commercially available workstation for reconstructions and 3D manipulation. Occlusion level and clot length, the quality of collateral vessels, and the patency of anterior and posterior communicating arteries were assessed. IV FDCT was performed successfully in all the cases and allowed for clot location and length visualization, assessment of communicating arteries patency, and evaluation of vessel collateral grade. Information obtained from this technique was considered useful for patients treated by endovascular approach. Retrospective review of the images by two independent readers was considered accurate and reproducible. IV FDCT technology provided accurate delineation of obstructed vessel segments in acute ischemic stroke disease. It gave a significant help in the interventional strategy. This new technology available in the operating room might provide a valuable tool in emerging endovascular stroke therapy. (orig.)

  19. Measurements of surgeons' exposure to ionizing radiation dose during intraoperative use of C-arm fluoroscopy.

    Science.gov (United States)

    Lee, Kisung; Lee, Kyoung Min; Park, Moon Seok; Lee, Boram; Kwon, Dae Gyu; Chung, Chin Youb

    2012-06-15

    Measurement of radiation dose from C-arm fluoroscopy during a simulated intraoperative use in spine surgery. OBJECTIVE.: To investigate scatter radiation doses to specific organs of surgeons during intraoperative use of C-arm fluoroscopy in spine surgery and to provide practical intraoperative guidelines. There have been studies that reported the radiation dose of C-arm fluoroscopy in various procedures. However, radiation doses to surgeons' specific organs during spine surgery have not been sufficiently examined, and the practical intraoperative radioprotective guidelines have not been suggested. Scatter radiation dose (air kerma rate) was measured during the use of a C-arm on an anthropomorphic chest phantom on an operating table. Then, a whole body anthropomorphic phantom was located besides the chest phantom to simulate a surgeon, and scatter radiation doses to specific organs (eye, thyroid, breast, and gonads) and direct radiation dose to the surgeon's hand were measured using 4 C-arm configurations (standard, inverted, translateral, and tube translateral). The effects of rotating the surgeon's head away from the patient and of a thyroid shield were also evaluated. Scatter radiation doses decreased as distance from the patient increased during C-arm fluoroscopy use. The standard and translateral C-arm configurations caused lower scatter doses to sensitive organs than inverted and tube translateral configurations. Scatter doses were highest for breast and lowest for gonads. The use of a thyroid shield and rotating the surgeon's head away from the patient reduced scatter radiation dose to the surgeon's thyroid and eyes. The direct radiation dose was at least 20 times greater than scatter doses to sensitive organs. The following factors could reduce radiation exposure during intraoperative use of C-arm; (1) distance from the patient, (2) C-arm configuration, (3) radioprotective equipments, (4) rotating the surgeons' eyes away from the patient, and (5) avoiding

  20. Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology

    Energy Technology Data Exchange (ETDEWEB)

    Struffert, Tobias; Hauer, Michael; Doerfler, Arnd [University of Erlangen-Nuremberg, Department of Neuroradiology, Erlangen (Germany); Banckwitz, Rosemarie; Koehler, Christoph [Siemens AG, Healthcare Sector, Forchheim (Germany); Royalty, Kevin [Siemens Medical Solutions, USA, Inc, Hoffman Estates, IL (United States); University of Wisconsin, Department of Biomedical Engineering and School of Medicine and Public Health, Madison, WI (United States)

    2014-06-15

    Flat-detector CT (FD-CT) is used for a variety of applications. Additionally, 3D rotational angiography (3D DSA) is used to supplement digital subtraction angiography (DSA) studies. The aim was to measure and compare the dose of (1) standard DSA and 3D DSA and (2) analogous FD-CT and multislice CT (MSCT) protocols. Using an anthropomorphic phantom, the effective dose to patients (according to ICRP 103) was measured on an MSCT and a flat-detector angiographic system using standard protocols as recommended by the manufacturer. (1) Evaluation of DSA and 3D DSA angiography protocols: ap.-lat. Standard/low-dose series 1/0.8 mSv, enlarged oblique projection 0.3 mSv, 3D DSA 0.9 mSv (limited coverage length 0.3 mSv). (2) Comparison of FD-CT and MSCT: brain parenchyma imaging 2.9 /1.4 mSv, perfusion imaging 2.3/4.2 mSv, temporal bone 0.2 /0.2 mSv, angiography 2.9/3.3 mSv, limited to the head using collimation 0.5/0.5 mSv. The effective dose for an FD-CT application depends on the application used. Using collimation for FD-CT applications, the dose may be reduced considerably. Due to the low dose of 3D DSA, we recommend using this technique to reduce the number of DSA series needed to identify working projections. (orig.)

  1. Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology

    International Nuclear Information System (INIS)

    Struffert, Tobias; Hauer, Michael; Doerfler, Arnd; Banckwitz, Rosemarie; Koehler, Christoph; Royalty, Kevin

    2014-01-01

    Flat-detector CT (FD-CT) is used for a variety of applications. Additionally, 3D rotational angiography (3D DSA) is used to supplement digital subtraction angiography (DSA) studies. The aim was to measure and compare the dose of (1) standard DSA and 3D DSA and (2) analogous FD-CT and multislice CT (MSCT) protocols. Using an anthropomorphic phantom, the effective dose to patients (according to ICRP 103) was measured on an MSCT and a flat-detector angiographic system using standard protocols as recommended by the manufacturer. (1) Evaluation of DSA and 3D DSA angiography protocols: ap.-lat. Standard/low-dose series 1/0.8 mSv, enlarged oblique projection 0.3 mSv, 3D DSA 0.9 mSv (limited coverage length 0.3 mSv). (2) Comparison of FD-CT and MSCT: brain parenchyma imaging 2.9 /1.4 mSv, perfusion imaging 2.3/4.2 mSv, temporal bone 0.2 /0.2 mSv, angiography 2.9/3.3 mSv, limited to the head using collimation 0.5/0.5 mSv. The effective dose for an FD-CT application depends on the application used. Using collimation for FD-CT applications, the dose may be reduced considerably. Due to the low dose of 3D DSA, we recommend using this technique to reduce the number of DSA series needed to identify working projections. (orig.)

  2. Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector

    International Nuclear Information System (INIS)

    Zhao, Z.; Gang, G. J.; Siewerdsen, J. H.

    2014-01-01

    Purpose: To investigate the effect of the number of projection views on image noise in cone-beam CT (CBCT) with a flat-panel detector. Methods: This fairly fundamental consideration in CBCT system design and operation was addressed experimentally (using a phantom presenting a uniform medium as well as statistically motivated “clutter”) and theoretically (using a cascaded systems model describing CBCT noise) to elucidate the contributing factors of quantum noise (σ Q ), electronic noise (σ E ), and view aliasing (σ view ). Analysis included investigation of the noise, noise-power spectrum, and modulation transfer function as a function of the number of projections (N proj ), dose (D tot ), and voxel size (b vox ). Results: The results reveal a nonmonotonic relationship between image noise andN proj at fixed total dose: for the CBCT system considered, noise decreased with increasing N proj due to reduction of view sampling effects in the regime N proj proj due to increased electronic noise. View sampling effects were shown to depend on the heterogeneity of the object in a direct analytical relationship to power-law anatomical clutter of the form κ/f  β —and a general model of individual noise components (σ Q , σ E , and σ view ) demonstrated agreement with measurements over a broad range in N proj , D tot , and b vox . Conclusions: The work elucidates fairly basic elements of CBCT noise in a manner that demonstrates the role of distinct noise components (viz., quantum, electronic, and view sampling noise). For configurations fairly typical of CBCT with a flat-panel detector (FPD), the analysis reveals a “sweet spot” (i.e., minimum noise) in the rangeN proj ∼ 250–350, nearly an order of magnitude lower in N proj than typical of multidetector CT, owing to the relatively high electronic noise in FPDs. The analysis explicitly relates view aliasing and quantum noise in a manner that includes aspects of the object (“clutter”) and imaging chain

  3. A compact high resolution flat panel PET detector based on the new 4-side buttable MPPC for biomedical applications.

    Science.gov (United States)

    Wang, Qiang; Wen, Jie; Ravindranath, Bosky; O'Sullivan, Andrew W; Catherall, David; Li, Ke; Wei, Shouyi; Komarov, Sergey; Tai, Yuan-Chuan

    2015-09-11

    Compact high-resolution panel detectors using virtual pinhole (VP) PET geometry can be inserted into existing clinical or pre-clinical PET systems to improve regional spatial resolution and sensitivity. Here we describe a compact panel PET detector built using the new Though Silicon Via (TSV) multi-pixel photon counters (MPPC) detector. This insert provides high spatial resolution and good timing performance for multiple bio-medical applications. Because the TSV MPPC design eliminates wire bonding and has a package dimension which is very close to the MPPC's active area, it is 4-side buttable. The custom designed MPPC array (based on Hamamatsu S12641-PA-50(x)) used in the prototype is composed of 4 × 4 TSV-MPPC cells with a 4.46 mm pitch in both directions. The detector module has 16 × 16 lutetium yttrium oxyorthosilicate (LYSO) crystal array, with each crystal measuring 0.92 × 0.92 × 3 mm 3 with 1.0 mm pitch. The outer diameter of the detector block is 16.8 × 16.8 mm 2 . Thirty-two such blocks will be arranged in a 4 × 8 array with 1 mm gaps to form a panel detector with detection area around 7 cm × 14 cm in the full-size detector. The flood histogram acquired with Ge-68 source showed excellent crystal separation capability with all 256 crystals clearly resolved. The detector module's mean, standard deviation, minimum (best) and maximum (worst) energy resolution were 10.19%, +/-0.68%, 8.36% and 13.45% FWHM, respectively. The measured coincidence time resolution between the block detector and a fast reference detector (around 200 ps single photon timing resolution) was 0.95 ns. When tested with Siemens Cardinal electronics the performance of the detector blocks remain consistent. These results demonstrate that the TSV-MPPC is a promising photon sensor for use in a flat panel PET insert composed of many high resolution compact detector modules.

  4. DQE of wireless digital detectors: Comparative performance with differing filtration schemes

    International Nuclear Information System (INIS)

    Samei, Ehsan; Murphy, Simon; Christianson, Olav

    2013-01-01

    Purpose: Wireless flat panel detectors are gaining increased usage in portable medical imaging. Two such detectors were evaluated and compared with a conventional flat-panel detector using the formalism of the International Electrotechnical Commission (IEC 62220-1) for measuring modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) using two different filtration schemes.Methods: Raw images were acquired for three image receptors (DRX-1C and DRX-1, Carestream Health; Inc., Pixium 4600, Trixell) using a radiographic system with a well-characterized output (Philips Super80 CP, Philips Healthcare). Free in-air exposures were measured using a calibrated radiation meter (Unfors Mult-O-Meter Type 407, Unfors Instruments AB). Additional aluminum filtration and a new alternative combined copper-aluminum filtration were used to conform the x ray output to IEC-specified beam quality definitions RQA5 and RQA9. Using the IEC 62220-1 formalism, each detector was evaluated at X N /2, X N , and 2X N , where the normal exposure level to the detector surface (X N ) was set to 8.73 μGy (1.0 mR). The prescribed edge test device was used to evaluate the MTF, while the NNPS was measured using uniform images. The DQE was then calculated from the MTF and NNPS and compared across detectors, exposures, and filtration schemes.Results: The three DR systems had largely comparable MTFs with DRX-1 demonstrating lower values above 1.0 cycles/mm. At each exposure, DRX-1C and Pixium detectors demonstrated better noise performance than that of DRX-1. Zero-frequency DQEs for DRX-1C, Pixium, and DRX-1 detectors were approximately 74%, 63%, and 38% for RQA5 and 50%, 42%, and 28% for RQA9, respectively.Conclusions: DRX-1C detector exhibited superior DQE performance compared to Pixium and DRX-1. In terms of filtration, the alternative filtration was found to provide comparable performance in terms of rank ordering of different detectors with

  5. Influence of a combined CT/C-arm system on periprocedural workflow and procedure times in mechanical thrombectomy

    Energy Technology Data Exchange (ETDEWEB)

    Pfaff, Johannes; Herweh, Christian; Pham, Mirko; Heiland, Sabine; Bendszus, Martin; Moehlenbruch, Markus Alfred [University of Heidelberg, Department of Neuroradiology, Heidelberg (Germany); Schoenenberger, Silvia; Nagel, Simon; Ringleb, Peter Arthur [University of Heidelberg, Department of Neurology, Heidelberg (Germany)

    2017-09-15

    To achieve the fastest possible workflow in ischaemic stroke, we developed a CT/C-arm system, which allows imaging and endovascular treatment on the same patient table. This prospective, monocentric trial was conducted between October 2014 and August 2016. Patients received stroke imaging and mechanical thrombectomy under general anaesthesia (GA) or conscious sedation (CS) using our combined setup comprising a CT-scanner and a mobile C-arm X-ray device. Primary endpoint was time between stroke imaging and groin puncture. We compared periprocedural workflow and procedure times with the literature and a matched patient cohort treated with a biplane angiographic system before installation of the CT/C-arm system. In 50 patients with acute ischaemic stroke due to large-vessel occlusion in the anterior circulation, comparable recanalization rates were achieved by using the CT/C-arm setup (TICI2b-3:CT/C-arm-GA: 85.7%; CT/C-arm-CS: 90.9%; Angiosuite: 78.6%; p = 0.269) without increasing periprocedural complications. Elimination of patient transport resulted in a significant reduction of the time between stroke imaging and groin puncture: median, min (IQR): CT/C-arm-GA: 43 (35-52); CT/C-arm-CS: 39 (28-49); Angiosuite: 64 (48-74); p < 0.0001. The combined CT/C-arm system allows comparable recanalization rates as a biplane angiographic system and accelerates the start of the endovascular stroke treatment. (orig.)

  6. Influence of a combined CT/C-arm system on periprocedural workflow and procedure times in mechanical thrombectomy

    International Nuclear Information System (INIS)

    Pfaff, Johannes; Herweh, Christian; Pham, Mirko; Heiland, Sabine; Bendszus, Martin; Moehlenbruch, Markus Alfred; Schoenenberger, Silvia; Nagel, Simon; Ringleb, Peter Arthur

    2017-01-01

    To achieve the fastest possible workflow in ischaemic stroke, we developed a CT/C-arm system, which allows imaging and endovascular treatment on the same patient table. This prospective, monocentric trial was conducted between October 2014 and August 2016. Patients received stroke imaging and mechanical thrombectomy under general anaesthesia (GA) or conscious sedation (CS) using our combined setup comprising a CT-scanner and a mobile C-arm X-ray device. Primary endpoint was time between stroke imaging and groin puncture. We compared periprocedural workflow and procedure times with the literature and a matched patient cohort treated with a biplane angiographic system before installation of the CT/C-arm system. In 50 patients with acute ischaemic stroke due to large-vessel occlusion in the anterior circulation, comparable recanalization rates were achieved by using the CT/C-arm setup (TICI2b-3:CT/C-arm-GA: 85.7%; CT/C-arm-CS: 90.9%; Angiosuite: 78.6%; p = 0.269) without increasing periprocedural complications. Elimination of patient transport resulted in a significant reduction of the time between stroke imaging and groin puncture: median, min (IQR): CT/C-arm-GA: 43 (35-52); CT/C-arm-CS: 39 (28-49); Angiosuite: 64 (48-74); p < 0.0001. The combined CT/C-arm system allows comparable recanalization rates as a biplane angiographic system and accelerates the start of the endovascular stroke treatment. (orig.)

  7. Angiography-based C-arm CT for the assessment of extrahepatic shunting before radioembolization

    International Nuclear Information System (INIS)

    Heusner, Till Alexander; Hahn, S.; Forsting, M.; Antoch, G.; Hamami, M.E.; Poeppel, T.; Bockisch, A.; Ertle, J.; Hilgard, P.

    2010-01-01

    Purpose: to retrospectively assess the accuracy of angiography-based C-arm CT for the detection of extrahepatic shunting before SIRT. Materials and methods: 30 patients (mean age: 64 ± 12 years) with hypervascularized hepatic tumors underwent hepatic angiography, coil embolization of gastrointestinal collaterals and 99mTc-macroaggregated albumin (MAA) SPECT/CT before SIRT. Before MAA injection via a microcatheter from the intended treatment position, an angiography and angiography-based C-arm CT (XperCT trademark, Philips Healthcare) were acquired. Angiographies and XperCT trademark were performed from 48 microcatheter positions followed by MAA injections and MAA-SPECT/CT. MAA-SPECT/CT served as the reference standard for determining the accuracy of hepatic arteriography and C-arm CT for the detection of extrahepatic shunting. Results: MAA-SPECT/CT revealed extrahepatic shunting in 5 patients (17%). Hepatic arteriography yielded a true negative in 22 (73%), a false negative in 5 (17%), and an unclear result in 3 patients (10%). C-arm CT yielded a true positive in 3 (10%), true negative in 24 (80%), false positive in 1 (3%), and false negative in 2 patients (7%). The specificity and the NPV of hepatic arteriography for the detection of extrahepatic shunting were 88% and 81%, respectively. For C-arm CT the sensitivity, specificity, PPV, NPV, and accuracy for the detection of extrahepatic shunting were 60%, 96%, 75%, 92%, and 90%, respectively. Conclusion: C-arm CT offers additional information to angiography when assessing SIRT patients for extrahepatic shunting. More accurate detection of extrahepatic shunting may optimize the workflow in SIRT preparations by avoiding unnecessary repeat angiographies. (orig.)

  8. Mobile C-arm cone-beam CT for guidance of spine surgery: Image quality, radiation dose, and integration with interventional guidance

    Energy Technology Data Exchange (ETDEWEB)

    Schafer, S.; Nithiananthan, S.; Mirota, D. J.; Uneri, A.; Stayman, J. W.; Zbijewski, W.; Schmidgunst, C.; Kleinszig, G.; Khanna, A. J.; Siewerdsen, J. H. [Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21202 (United States); Department of Computer Science, Johns Hopkins University, Baltimore, Maryland 21218 (United States); Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21202 (United States); Siemens Healthcare XP Division, Erlangen (Germany); Department of Orthopaedic Surgery, Johns Hopkins University, Baltimore, Maryland 21239 (United States); Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21202 and Department of Computer Science, Johns Hopkins University, Baltimore, Maryland 21218 (United States)

    2011-08-15

    Purpose: A flat-panel detector based mobile isocentric C-arm for cone-beam CT (CBCT) has been developed to allow intraoperative 3D imaging with sub-millimeter spatial resolution and soft-tissue visibility. Image quality and radiation dose were evaluated in spinal surgery, commonly relying on lower-performance image intensifier based mobile C-arms. Scan protocols were developed for task-specific imaging at minimum dose, in-room exposure was evaluated, and integration of the imaging system with a surgical guidance system was demonstrated in preclinical studies of minimally invasive spine surgery. Methods: Radiation dose was assessed as a function of kilovolt (peak) (80-120 kVp) and milliampere second using thoracic and lumbar spine dosimetry phantoms. In-room radiation exposure was measured throughout the operating room for various CBCT scan protocols. Image quality was assessed using tissue-equivalent inserts in chest and abdomen phantoms to evaluate bone and soft-tissue contrast-to-noise ratio as a function of dose, and task-specific protocols (i.e., visualization of bone or soft-tissues) were defined. Results were applied in preclinical studies using a cadaveric torso simulating minimally invasive, transpedicular surgery. Results: Task-specific CBCT protocols identified include: thoracic bone visualization (100 kVp; 60 mAs; 1.8 mGy); lumbar bone visualization (100 kVp; 130 mAs; 3.2 mGy); thoracic soft-tissue visualization (100 kVp; 230 mAs; 4.3 mGy); and lumbar soft-tissue visualization (120 kVp; 460 mAs; 10.6 mGy) - each at (0.3 x 0.3 x 0.9 mm{sup 3}) voxel size. Alternative lower-dose, lower-resolution soft-tissue visualization protocols were identified (100 kVp; 230 mAs; 5.1 mGy) for the lumbar region at (0.3 x 0.3 x 1.5 mm{sup 3}) voxel size. Half-scan orbit of the C-arm (x-ray tube traversing under the table) was dosimetrically advantageous (prepatient attenuation) with a nonuniform dose distribution ({approx}2 x higher at the entrance side than at isocenter

  9. The one-armed ATLAS Forward Proton detector

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00372192; Lange, Joern

    The ATLAS experiment at the European Laboratory for Particle Physics (CERN), Geneva, has been taking data successfully since the Large Hadron Collider (LHC) accelerator started operations in 2010. Since then, it has been generating proton-proton collisions to study the frontiers of particle physics, at a centre of mass energy of 7-8 TeV first and, more recently, 13 TeV. However, the experiment is in constant evolution: detectors ageing due to radiation damage, increasing collision rates and pile-up, and new scientific objectives often require upgrades of the ATLAS detectors. These ever-growing challenges motivate the continued research and development of new detector technologies. To enhance the physics search of the experiment the ATLAS collaboration recently added a forward detector to identify intact protons that emerge from LHC collisions at very shallow angles. The ATLAS Forward Proton (AFP) detector enables the identification of diffractive processes and, ultimately, of central exclusive events, thus al...

  10. Marker detection evaluation by phantom and cadaver experiments for C-arm pose estimation pattern

    Science.gov (United States)

    Steger, Teena; Hoßbach, Martin; Wesarg, Stefan

    2013-03-01

    C-arm fluoroscopy is used for guidance during several clinical exams, e.g. in bronchoscopy to locate the bronchoscope inside the airways. Unfortunately, these images provide only 2D information. However, if the C-arm pose is known, it can be used to overlay the intrainterventional fluoroscopy images with 3D visualizations of airways, acquired from preinterventional CT images. Thus, the physician's view is enhanced and localization of the instrument at the correct position inside the bronchial tree is facilitated. We present a novel method for C-arm pose estimation introducing a marker-based pattern, which is placed on the patient table. The steel markers form a pattern, allowing to deduce the C-arm pose by use of the projective invariant cross-ratio. Simulations show that the C-arm pose estimation is reliable and accurate for translations inside an imaging area of 30 cm x 50 cm and rotations up to 30°. Mean error values are 0.33 mm in 3D space and 0.48 px in the 2D imaging plane. First tests on C-arm images resulted in similarly compelling accuracy values and high reliability in an imaging area of 30 cm x 42.5 cm. Even in the presence of interfering structures, tested both with anatomy phantoms and a turkey cadaver, high success rates over 90% and fully satisfying execution times below 4 sec for 1024 px × 1024 px images could be achieved.

  11. Development of high quantum efficiency, flat panel, thick detectors for megavoltage x-ray imaging: An experimental study of a single-pixel prototype

    International Nuclear Information System (INIS)

    Mei, X.; Pang, G.

    2005-01-01

    Our overall goal is to develop a new generation of electronic portal imaging devices (EPIDs) with a quantum efficiency (QE) more than an order of magnitude higher and a spatial resolution equivalent to that of EPIDs currently used for portal imaging. A novel design of such a high QE flat-panel based EPID was introduced recently and its feasibility was investigated theoretically [see Pang and Rowlands, Med. Phys. 31, 3004 (2004)]. In this work, we constructed a prototype single-pixel detector based on the novel design. Some fundamental imaging properties including the QE, spatial resolution, and sensitivity of the prototype detector were measured with a 6 MV beam. It has been shown that the experimental results agree well with theoretical predictions and further development based on the novel design including the construction of a prototype area detector is warranted

  12. Clinical evaluation of digital radiography based on a large-area cesium iodide-amorphous silicon flat-panel detector compared with screen-film radiography for skeletal system and abdomen

    International Nuclear Information System (INIS)

    Okamura, Terue; Tanaka, Saori; Koyama, Koichi; Norihumi, Nishida; Daikokuya, Hideo; Matsuoka, Toshiyuki; Yamada, Ryusaku; Kishimoto, Kenji; Hatagawa, Masakatsu; Kudoh, Hiroaki

    2002-01-01

    The aim of this clinical study was to compare the image quality of digital radiography using the new digital Bucky system based on a flat-panel detector with that of a conventional screen-film system for the skeletal structure and the abdomen. Fifty patients were examined using digital radiography with a flat-panel detector and screen-film systems, 25 for the skeletal structures and 25 for the abdomen. Six radiologists judged each paired image acquired under the same exposure parameters concerning three observation items for the bone and six items for the abdomen. Digital radiographic images for the bone were evaluated to be similar to screen-film images at the mean of 42.2%, to be superior at 50.2%, and to be inferior at 7.6%. Digital radiographic images for the abdomen were judged to be similar to screen-film images at the mean of 43.4%, superior at 52.4%, and inferior at 4.2%; thus, digital radiographic images were estimated to be either similar as or superior to screen-film images at over 92% for the bone and abdomen. On the statistical analysis, digital radiographic images were also judged to be preferred significantly in the most items for the bone and abdomen. In conclusion, the image quality of digital radiography with a flat-panel detector was superior to that of a screen-film system under the same exposure parameters, suggesting that dose reduction is possible with digital radiography. (orig.)

  13. A Fast, Accurate and Easy to Implement Method for Pose Recognition of an Intramedullary Nail using a Tracked C-arm

    Directory of Open Access Journals (Sweden)

    H. Esfandiari

    2014-06-01

    Full Text Available A C-arm is a mobile X-ray device that is frequently used during orthopaedic surgeries. It consists of a semi-circular, arc-shaped arm that holds an X-ray transmitter at one end and an X-ray detector at the other. Intramedullary nail (IM nail fixation is a popular orthopaedic surgery in which a metallic rod is placed into the patient's fractured bone (femur or tibia and fixed using metal screws. The main challenge of IM-nail fixation surgery is to achieve the X-ray shot in which the distal holes of the IM nail appear as circles (desired view so that the surgeon can easily insert the screws. Although C-arm X-ray devices are routinely used in IM-nail fixation surgeries, the surgeons or radiation technologists (rad-techs usually use it in a trial-and-error manner. This method raises both radiation exposure and surgery time. In this study, we have designed and developed an IM-nail distal locking navigation technique that leads to more accurate and faster screw placement with a lower radiation dose and a minimum number of added steps to the operation to make it more accepted within the orthopaedic community. The specific purpose of this study was to develop and validate an automated technique for identifying the current pose of the IM nail relative to the C-arm. An accuracy assessment was performed to test the reliability of the navigation results. Translational accuracy was demonstrated to be better than 1 mm, roll and pitch rotations better than 2° and yaw rotational accuracy better than 2–5° depending on the separate angle. Computation time was less than 3.5 seconds.

  14. Indirect flat-panel detector with avalanche gain: Fundamental feasibility investigation for SHARP-AMFPI (scintillator HARP active matrix flat panel imager)

    International Nuclear Information System (INIS)

    Zhao Wei; Li Dan; Reznik, Alla; Lui, B.J.M.; Hunt, D.C.; Rowlands, J.A.; Ohkawa, Yuji; Tanioka, Kenkichi

    2005-01-01

    An indirect flat-panel imager (FPI) with avalanche gain is being investigated for low-dose x-ray imaging. It is made by optically coupling a structured x-ray scintillator CsI(Tl) to an amorphous selenium (a-Se) avalanche photoconductor called HARP (high-gain avalanche rushing photoconductor). The final electronic image is read out using an active matrix array of thin film transistors (TFT). We call the proposed detector SHARP-AMFPI (scintillator HARP active matrix flat panel imager). The advantage of the SHARP-AMFPI is its programmable gain, which can be turned on during low dose fluoroscopy to overcome electronic noise, and turned off during high dose radiography to avoid pixel saturation. The purpose of this paper is to investigate the important design considerations for SHARP-AMFPI such as avalanche gain, which depends on both the thickness d Se and the applied electric field E Se of the HARP layer. To determine the optimal design parameter and operational conditions for HARP, we measured the E Se dependence of both avalanche gain and optical quantum efficiency of an 8 μm HARP layer. The results were used in a physical model of HARP as well as a linear cascaded model of the FPI to determine the following x-ray imaging properties in both the avalanche and nonavalanche modes as a function of E Se : (1) total gain (which is the product of avalanche gain and optical quantum efficiency); (2) linearity; (3) dynamic range; (4) gain nonuniformity resulting from thickness nonuniformity; and (5) effects of direct x-ray interaction in HARP. Our results showed that a HARP layer thickness of 8 μm can provide adequate avalanche gain and sufficient dynamic range for x-ray imaging applications to permit quantum limited operation over the range of exposures needed for radiography and fluoroscopy

  15. Intra-operative adjustment of standard planes in C-arm CT image data.

    Science.gov (United States)

    Brehler, Michael; Görres, Joseph; Franke, Jochen; Barth, Karl; Vetter, Sven Y; Grützner, Paul A; Meinzer, Hans-Peter; Wolf, Ivo; Nabers, Diana

    2016-03-01

    With the help of an intra-operative mobile C-arm CT, medical interventions can be verified and corrected, avoiding the need for a post-operative CT and a second intervention. An exact adjustment of standard plane positions is necessary for the best possible assessment of the anatomical regions of interest but the mobility of the C-arm causes the need for a time-consuming manual adjustment. In this article, we present an automatic plane adjustment at the example of calcaneal fractures. We developed two feature detection methods (2D and pseudo-3D) based on SURF key points and also transferred the SURF approach to 3D. Combined with an atlas-based registration, our algorithm adjusts the standard planes of the calcaneal C-arm images automatically. The robustness of the algorithms is evaluated using a clinical data set. Additionally, we tested the algorithm's performance for two registration approaches, two resolutions of C-arm images and two methods for metal artifact reduction. For the feature extraction, the novel 3D-SURF approach performs best. As expected, a higher resolution ([Formula: see text] voxel) leads also to more robust feature points and is therefore slightly better than the [Formula: see text] voxel images (standard setting of device). Our comparison of two different artifact reduction methods and the complete removal of metal in the images shows that our approach is highly robust against artifacts and the number and position of metal implants. By introducing our fast algorithmic processing pipeline, we developed the first steps for a fully automatic assistance system for the assessment of C-arm CT images.

  16. Utility of intraoperative diagnostic C-arm angiography for management of high grade subarachnoid hemorrhage

    Directory of Open Access Journals (Sweden)

    Zhikui Wei

    2015-06-01

    Full Text Available The accurate and efficient localization of underlying vascular lesions is crucial for prompt and definitive treatment of subarachnoid hemorrhage (SAH. To demonstrate the utility and feasibility of intraoperative C-arm angiography in cerebrovascular emergencies, we report five cases of high grade SAH and/or intracerebral hemorrhage (ICH where intraoperative diagnostic C-arm angiography was safely and effectively utilized. Initial evaluations of all patients included a non-contrast head CT scan, which was followed by urgent decompressive hemicraniectomy as a life-saving measure in the presence of markedly elevated intracranial pressure. Further diagnostic evaluations were performed intraoperatively using a multi-purpose C-arm angiography system. The C-arm angiography findings greatly aided the intraoperative planning and led to definitive treatments in four cases of SAH by elucidating the underlying neurovascular lesions. With this treatment strategy, two of the patients made moderately good recoveries from their SAH and/or ICH with a Glasgow outcome score (GOS of 4. Three of the patients expired despite maximal therapy mostly due to unfavorable presenting grade. These results suggest that C-arm angiography is a reasonable diagnostic and surgical planning tool for selected patients with high grade diffuse SAH who require immediate decompression.

  17. A Scott bench with ergonomic thorax stabilisation pad improves body posture during preacher arm curl exercise.

    Science.gov (United States)

    Biscarini, Andrea; Benvenuti, Paolo; Busti, Daniele; Zanuso, Silvano

    2016-05-01

    We assessed whether the use of an ergonomic thorax stabilisation pad, during the preacher arm curl exercise, could significantly reduce the excessive shoulder protraction and thoracic kyphosis induced by the standard flat pad built into the existing preacher arm curl equipment. A 3D motion capture system and inclinometers were used to measure shoulder protraction and thoracic kyphosis in 15 subjects performing preacher arm curl with a plate-loaded machine provided with the standard flat pad. The same measures were repeated after replacing the flat pad with a new ergonomic pad, specifically designed to accommodate the thorax profile and improve body posture. Pad replacement significantly (p ergonomic pad may potentially allow a more effective training, prevent musculoskeletal discomfort and reduce the risk of injury. Practitioner summary: We have designed an ergonomic thorax stabilisation pad for the preacher arm curl exercise. The new ergonomic pad improves the poor posture conditions induced by the standard flat pad and may potentially allow a more effective training, prevent musculoskeletal discomfort, improve the breathing function and reduce the risk of injury.

  18. UVIS Flat Field Uniformity

    Science.gov (United States)

    Quijano, Jessica Kim

    2009-07-01

    The stability and uniformity of the low-frequency flat fields {L-flat} of the UVIS detector will be assessed by using multiple-pointing observations of the globular clusters 47 Tucanae {NGC104} and Omega Centauri {NGC5139}, thus imaging moderately dense stellar fields. By placing the same star over different portions of the detector and measuring relative changes in its brightness, it will be possible to determine local variations in the response of the UVIS detector. Based on previous experience with STIS and ACS, it is deemed that a total of 9 different pointings will suffice to provide adequate characterization of the flat field stability in any given band. For each filter to be tested, the baseline consists of 9 pointings in a 3X3 box pattern with dither steps of about 25% of the FOV, or 40.5", in either the x or y direction {useful also for CTE measurements, if needed in the future}. During SMOV, the complement of filters to be tested is limited to the following 6 filters: F225W, F275W, F336W, for Omega Cen, and F438W, F606W, and F814W for 47 Tuc. Three long exposures for each target are arranged such that the initial dither position is observed with the appropriate filters for that target within one orbit at a single pointing, so that filter-to-filter differences in the observed star positions can be checked. In addition to the 9 baseline exposures, two sets of short exposures will be taken:a} one short exposure will be taken of OmegaCen with each of the visible filters {F438W, F606W and F814W} in order to check the geometric distortion solution to be obtained with the data from proposal 11444;b} for each target, a single short exposure will be taken with each filter to facilitate the study of the PSF as a function of position on the detector by providing unsaturated images of sparsely-spaced bright stars.This proposal corresponds to Activity Description ID WF39. It should execute only after the following proposal has executed:WF21 - 11434

  19. Quantification of the gravity-dependent change in the C-arm image center for image compensation in fluoroscopic spinal neuronavigation.

    Science.gov (United States)

    Hariri, S; Abbasi, H R; Chin, S; Steinberg, G; Shahidi, R

    2001-01-01

    In the quest to develop a viable, frameless spinal navigation system, many researchers are utilizing the C-arm fluoroscope. However, there is a significant problem with the C-arm that must be quantified: the gravity-dependent sag effect resulting from the geometry of the C-arm and aggravated by the inequity of weight at each end of the C-arm. This study quantified the C-arm sag effect, giving researchers the protocol and data needed to develop a program that accounts for this distortion. The development of spinal navigation algorithms that account for the C-arm sag effect should produce a more accurate spinal navigation system.

  20. Studies on Flat Sandwich-type Self-Powered Detectors for Flux Measurements in ITER Test Blanket Modules

    Science.gov (United States)

    Raj, Prasoon; Angelone, Maurizio; Döring, Toralf; Eberhardt, Klaus; Fischer, Ulrich; Klix, Axel; Schwengner, Ronald

    2018-01-01

    Neutron and gamma flux measurements in designated positions in the test blanket modules (TBM) of ITER will be important tasks during ITER's campaigns. As part of the ongoing task on development of nuclear instrumentation for application in European ITER TBMs, experimental investigations on self-powered detectors (SPD) are undertaken. This paper reports the findings of neutron and photon irradiation tests performed with a test SPD in flat sandwich-like geometry. Whereas both neutrons and gammas can be detected with appropriate optimization of geometries, materials and sizes of the components, the present sandwich-like design is more sensitive to gammas than 14 MeV neutrons. Range of SPD current signals achievable under TBM conditions are predicted based on the SPD sensitivities measured in this work.

  1. Gain and offset calibration reduces variation in exposure-dependent SNR among systems with identical digital flat-panel detectors.

    Science.gov (United States)

    Willis, Charles E; Vinogradskiy, Yevgeniy Y; Lofton, Brad K; White, R Allen

    2011-07-01

    The conditions under which vendor performance criteria for digital radiography systems are obtained do not adequately simulate the conditions of actual clinical imaging with respect to radiographic technique factors, scatter production, and scatter control. Therefore, the relationship between performance under ideal conditions and performance in clinical practice remains unclear. Using data from a large complement of systems in clinical use, the authors sought to develop a method to establish expected performance criteria for digital flat-panel radiography systems with respect to signal-to-noise ratio (SNR) versus detector exposure under clinical conditions for thoracic imaging. The authors made radiographic exposures of a patient-equivalent chest phantom at 125 kVp and 180 cm source-to-image distance. The mAs value was modified to produce exposures above and below the mAs delivered by automatic exposure control. Exposures measured free-in-air were corrected to the imaging plane by the inverse square law, by the attenuation factor of the phantom, and by the Bucky factor of the grid for the phantom, geometry, and kilovolt peak. SNR was evaluated as the ratio of the mean to the standard deviation (SD) of a region of interest automatically selected in the center of each unprocessed image. Data were acquired from 18 systems, 14 of which were tested both before and after gain and offset calibration. SNR as a function of detector exposure was interpolated using a double logarithmic function to stratify the data into groups of 0.2, 0.5, 1.0, 2.0, and 5.0 mR exposure (1.8, 4.5, 9.0, 18, and 45 microGy air KERMA) to the detector. The mean SNR at each exposure interval after calibration exhibited linear dependence on the mean SNR before calibration (r2=0.9999). The dependence was greater than unity (m = 1.101 +/- 0.006), and the difference from unity was statistically significant (p grid replacement. The nonconformant behavior of the other system was corrected by replacing

  2. Imaging responses of on-site CsI and Gd2O2S flat-panel detectors: Dependence on the tube voltage

    Science.gov (United States)

    Jeon, Hosang; Chung, Myung Jin; Youn, Seungman; Nam, Jiho; Lee, Jayoung; Park, Dahl; Kim, Wontaek; Ki, Yongkan; Kim, Ho Kyung

    2015-07-01

    One of the emerging issues in radiography is low-dose imaging to minimize patient's exposure. The scintillating materials employed in most indirect flat-panel detectors show a drastic change of X-ray photon absorption efficiency around their K-edge energies that consequently affects image quality. Using various tube voltages, we investigated the imaging performance of most popular scintillators: cesium iodide (CsI) and gadolinium oxysulfide (Gd2O2S). The integrated detective quantum efficiencies (iDQE) of four detectors installed in the same hospital were evaluated according to the standardized procedure IEC 62220-1 at tube voltages of 40 - 120 kVp. The iDQE values of the Gd2O2S detectors were normalized by those of CsI detectors to exclude the effects of image postprocessing. The contrast-to-noise ratios (CNR) were also evaluated by using an anthropomorphic chest phantom. The iDQE of the CsI detector outperformed that of the Gd2O2S detector over all tube voltages. Moreover, we noted that the iDQE of the Gd2O2S detectors quickly rolled off with decreasing tube voltage under 70 kVp. The CNRs of the two scintillators were similar at 120 kVp. At 60 kVp, however, the CNR of Gd2O2S was about half that of CsI. Compared to the Gd2O2S detectors, variations in the DQE performance of the CsI detectors were relatively immune to variations in the applied tube voltages. Therefore, we claim that Gd2O2S detectors are inappropriate for use in low-tube-voltage imaging (e.g., extremities and pediatrics) with low patient exposure.

  3. DQE of wireless digital detectors: Comparative performance with differing filtration schemes

    Energy Technology Data Exchange (ETDEWEB)

    Samei, Ehsan [Carl E. Ravin Advanced Imaging Laboratories, Departments of Radiology, Biomedical Engineering, Physics, and Electrical and Computer Engineering, Medical Physics Graduate Program, Duke University Medical Center, Durham, North Carolina 27705 (United States); Murphy, Simon; Christianson, Olav [Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 (United States)

    2013-08-15

    Purpose: Wireless flat panel detectors are gaining increased usage in portable medical imaging. Two such detectors were evaluated and compared with a conventional flat-panel detector using the formalism of the International Electrotechnical Commission (IEC 62220-1) for measuring modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) using two different filtration schemes.Methods: Raw images were acquired for three image receptors (DRX-1C and DRX-1, Carestream Health; Inc., Pixium 4600, Trixell) using a radiographic system with a well-characterized output (Philips Super80 CP, Philips Healthcare). Free in-air exposures were measured using a calibrated radiation meter (Unfors Mult-O-Meter Type 407, Unfors Instruments AB). Additional aluminum filtration and a new alternative combined copper-aluminum filtration were used to conform the x ray output to IEC-specified beam quality definitions RQA5 and RQA9. Using the IEC 62220-1 formalism, each detector was evaluated at X{sub N}/2, X{sub N}, and 2X{sub N}, where the normal exposure level to the detector surface (X{sub N}) was set to 8.73 μGy (1.0 mR). The prescribed edge test device was used to evaluate the MTF, while the NNPS was measured using uniform images. The DQE was then calculated from the MTF and NNPS and compared across detectors, exposures, and filtration schemes.Results: The three DR systems had largely comparable MTFs with DRX-1 demonstrating lower values above 1.0 cycles/mm. At each exposure, DRX-1C and Pixium detectors demonstrated better noise performance than that of DRX-1. Zero-frequency DQEs for DRX-1C, Pixium, and DRX-1 detectors were approximately 74%, 63%, and 38% for RQA5 and 50%, 42%, and 28% for RQA9, respectively.Conclusions: DRX-1C detector exhibited superior DQE performance compared to Pixium and DRX-1. In terms of filtration, the alternative filtration was found to provide comparable performance in terms of rank ordering of different

  4. Comparison of imaging properties of direct-type and indirect-type of flat-panel detector

    International Nuclear Information System (INIS)

    Matsumoto, Masao; Suekane, Koji; Ichimaru, Yasunobu; Ogata, Yuji; Inamura, Kiyonari; Kanai, Kouzou; Kanamori, Hitoshi

    2002-01-01

    A Flat-Panel Detector (FPD) has many advantages such as eliminating cassette handling and being able to display a preview image immediately in addition to the digital image processing and the networking. Thus, the FPD has ability to innovate the radiology department. We measured and evaluated the digital and over-all imaging properties (characteristic curves, modulation Transfer Functions, Wiener spectra and Noise Equivalent Quanta (NEQ) for the direct-type and indirect-type of FPD. The pre-sampling and overall NEQ of the indirect-type of FPD were better than the NEQ of the direct-type of FPD at lower spatial frequencies, but were worse at higher spatial frequencies. The FPD can take image data at real-time and be easy to digitalize. From these results, Screen/Film system and Computed Radiography system will be replaced by the FPD system, together with diffusion of CAD, cone beam Computed Tomography (CT) system and open-type Magnetic Resonance Imagining (MRI) system. (T. Tanaka)

  5. X-ray detector array

    International Nuclear Information System (INIS)

    Houston, J.M.

    1980-01-01

    The object of the invention (an ionization chamber X-ray detector array for use with high speed computerised tomographic imaging apparatus) is to reduce the time required to produce a tomographic image. The detector array described determines the distribution of X-ray intensities in one or more flat, coplanar X-ray beams. It comprises three flat anode sheets parallel to the X-ray beam, a plurality of rod-like cathodes between the anodes, a detector gas between the electrodes and a means for applying a potential between the electrodes. Each of the X-ray sources is collimated to give a narrow, planar section of X-ray photons. Sets of X-ray sources in the array are pulsed simultaneously to obtain X-ray transmission data for tomographic image reconstruction. (U.K.)

  6. The PHOBOS detector at RHIC

    Science.gov (United States)

    Back, B. B.; Baker, M. D.; Barton, D. S.; Basilev, S.; Baum, R.; Betts, R. R.; Białas, A.; Bindel, R.; Bogucki, W.; Budzanowski, A.; Busza, W.; Carroll, A.; Ceglia, M.; Chang, Y.-H.; Chen, A. E.; Coghen, T.; Connor, C.; Czyż, W.; Dabrowski, B.; Decowski, M. P.; Despet, M.; Fita, P.; Fitch, J.; Friedl, M.; Gałuszka, K.; Ganz, R.; Garcia, E.; George, N.; Godlewski, J.; Gomes, C.; Griesmayer, E.; Gulbrandsen, K.; Gushue, S.; Halik, J.; Halliwell, C.; Haridas, P.; Hayes, A.; Heintzelman, G. A.; Henderson, C.; Hollis, R.; Hołyński, R.; Hofman, D.; Holzman, B.; Johnson, E.; Kane, J.; Katzy, J.; Kita, W.; Kotuła, J.; Kraner, H.; Kucewicz, W.; Kulinich, P.; Law, C.; Lemler, M.; Ligocki, J.; Lin, W. T.; Manly, S.; McLeod, D.; Michałowski, J.; Mignerey, A.; Mülmenstädt, J.; Neal, M.; Nouicer, R.; Olszewski, A.; Pak, R.; Park, I. C.; Patel, M.; Pernegger, H.; Plesko, M.; Reed, C.; Remsberg, L. P.; Reuter, M.; Roland, C.; Roland, G.; Ross, D.; Rosenberg, L.; Ryan, J.; Sanzgiri, A.; Sarin, P.; Sawicki, P.; Scaduto, J.; Shea, J.; Sinacore, J.; Skulski, W.; Steadman, S. G.; Stephans, G. S. F.; Steinberg, P.; Straczek, A.; Stodulski, M.; Strek, M.; Stopa, Z.; Sukhanov, A.; Surowiecka, K.; Tang, J.-L.; Teng, R.; Trzupek, A.; Vale, C.; van Nieuwenhuizen, G. J.; Verdier, R.; Wadsworth, B.; Wolfs, F. L. H.; Wosiek, B.; Woźniak, K.; Wuosmaa, A. H.; Wysłouch, B.; Zalewski, K.; Żychowski, P.; Phobos Collaboration

    2003-03-01

    This manuscript contains a detailed description of the PHOBOS experiment as it is configured for the Year 2001 running period. It is capable of detecting charged particles over the full solid angle using a multiplicity detector and measuring identified charged particles near mid-rapidity in two spectrometer arms with opposite magnetic fields. Both of these components utilize silicon pad detectors for charged particle detection. The minimization of material between the collision vertex and the first layers of silicon detectors allows for the detection of charged particles with very low transverse momenta, which is a unique feature of the PHOBOS experiment. Additional detectors include a time-of-flight wall which extends the particle identification range for one spectrometer arm, as well as sets of scintillator paddle and Cherenkov detector arrays for event triggering and centrality selection.

  7. Cone-Beam CT with Flat-Panel-Detector Digital Angiography System: Early Experience in Abdominal Interventional Procedures

    International Nuclear Information System (INIS)

    Hirota, Shozo; Nakao, Norio; Yamamoto, Satoshi; Kobayashi, Kaoru; Maeda, Hiroaki; Ishikura, Reiichi; Miura, Koui; Sakamoto, Kiyoshi; Ueda, Ken; Baba, Rika

    2006-01-01

    We developed a cone-beam computed tomography (CBCT) system equipped with a large flat-panel detector. Data obtained by 200 o rotation imaging are reconstructed by means of CBCT to generate three-dimensional images. We report the use of CBCT angiography using CBCT in 10 patients with 8 liver malignancies and 2 hypersplenisms during abdominal interventional procedures. CBCT was very useful for interventional radiologists to confirm a perfusion area of the artery catheter wedged on CT by injection of contrast media through the catheter tip, although the image quality was slightly degraded, scoring as 2.60 on average by streak artifacts. CBCT is space-saving because it does not require a CT system with a gantry, and it is also time-saving because it does not require the transfer of patients

  8. Focal shift and faculae dimension of focused flat beam propagating in turbulent atmosphere

    International Nuclear Information System (INIS)

    Zhang Jianzhu; Li Youkuan; Zhang Feizhou; An Jianzhu

    2011-01-01

    Through theoretic analysis and numerical simulation,the focal shift of a focused flat beam propagating in turbulent atmosphere is studied. When a focused flat beam propagates in turbulent atmosphere, the effect of turbulence will induce the focal spot to move toward the transmitter. The turbulence is stronger and the diameter of transmitter is smaller, the measure of focal shift is larger. When adjusting the focus of transmitter and letting the focal spot of beam locate on detector, the laser intensity received by detector is not the strongest. The laser intensity will be the strongest if the focus of transmitter equals to the distance from transmitter to detector. (authors)

  9. Isolated effects of peripheral arm and central body cooling on arm performance.

    Science.gov (United States)

    Giesbrecht, G G; Wu, M P; White, M D; Johnston, C E; Bristow, G K

    1995-10-01

    Whole body cooling impairs manual arm performance. The independent contributions of local (peripheral) and/or whole body (central) cooling are not known. Therefore, a protocol was developed in which the arm and the rest of the body could be independently cooled. Biceps temperature (Tmus), at a depth of 20 mm, and esophageal temperature (Tes) were measured. Six subjects were immersed to the clavicles in a tank (body tank) of water under 3 conditions: 1) cold body-cold arm (CB-CA); 2) warm body-cold arm (WB-CA); and 3) cold body-warm arm (CB-WA). In the latter two conditions, subjects placed their dominant arm in a separate (arm) tank. Water temperature (Tw) in each tank was independently controlled. In conditions requiring cold body and/or cold arm, Tw in the appropriate tanks was 8 degrees C. In conditions requiring warm body and/or warm arm, Tw in the appropriate tanks was adjusted between 29 and 38 degrees C to maintain body/arm temperature at baseline values. A battery of 6 tests, requiring fine or gross motor movements, were performed immediately before immersion and after 15, 45, and 70 minutes of immersion. In CB-CA, Tes decreased from an average of 37.2 to 35.6 degrees C and Tmus decreased from 34.6 to 22.0 degrees C. In WB-CA, Tmus decreased to 18.1 degrees C (Tes = 37.1 degrees C), and in CB-WA, Tes decreased to 35.8 degrees C (Tmus = 34.5 degrees C). By the end of immersion, there were significant decrements (43-85%) in the performance of all tests in CB-CA and WB-CA (p body and/or the arm elicits large decrements in finger, hand and arm performance. The decrements are due almost entirely to the local effects of arm tissue cooling.

  10. Image features for misalignment correction in medical flat-detector CT

    International Nuclear Information System (INIS)

    Wicklein, Julia; Kunze, Holger; Kalender, Willi A.; Kyriakou, Yiannis

    2012-01-01

    Purpose: Misalignment artifacts are a serious problem in medical flat-detector computed tomography. Generally, the geometrical parameters, which are essential for reconstruction, are provided by preceding calibration routines. These procedures are time consuming and the later use of stored parameters is sensitive toward external impacts or patient movement. The method of choice in a clinical environment would be a markerless online-calibration procedure that allows flexible scan trajectories and simultaneously corrects misalignment and motion artifacts during the reconstruction process. Therefore, different image features were evaluated according to their capability of quantifying misalignment. Methods: Projections of the FORBILD head and thorax phantoms were simulated. Additionally, acquisitions of a head phantom and patient data were used for evaluation. For the reconstruction different sources and magnitudes of misalignment were introduced in the geometry description. The resulting volumes were analyzed by entropy (based on the gray-level histogram), total variation, Gabor filter texture features, Haralick co-occurrence features, and Tamura texture features. The feature results were compared to the back-projection mismatch of the disturbed geometry. Results: The evaluations demonstrate the ability of several well-established image features to classify misalignment. The authors elaborated the particular suitability of the gray-level histogram-based entropy on identifying misalignment artifacts, after applying an appropriate window level (bone window). Conclusions: Some of the proposed feature extraction algorithms show a strong correlation with the misalignment level. Especially, entropy-based methods showed very good correspondence, with the best of these being the type that uses the gray-level histogram for calculation. This makes it a suitable image feature for online-calibration.

  11. Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging.

    Science.gov (United States)

    Tanaka, Rie

    2016-07-01

    Dynamic chest radiography is a flat-panel detector (FPD)-based functional X-ray imaging, which is performed as an additional examination in chest radiography. The large field of view (FOV) of FPDs permits real-time observation of the entire lungs and simultaneous right-and-left evaluation of diaphragm kinetics. Most importantly, dynamic chest radiography provides pulmonary ventilation and circulation findings as slight changes in pixel value even without the use of contrast media; the interpretation is challenging and crucial for a better understanding of pulmonary function. The basic concept was proposed in the 1980s; however, it was not realized until the 2010s because of technical limitations. Dynamic FPDs and advanced digital image processing played a key role for clinical application of dynamic chest radiography. Pulmonary ventilation and circulation can be quantified and visualized for the diagnosis of pulmonary diseases. Dynamic chest radiography can be deployed as a simple and rapid means of functional imaging in both routine and emergency medicine. Here, we focus on the evaluation of pulmonary ventilation and circulation. This review article describes the basic mechanism of imaging findings according to pulmonary/circulation physiology, followed by imaging procedures, analysis method, and diagnostic performance of dynamic chest radiography.

  12. Neutron damage of silicon detectors at 20 deg C

    International Nuclear Information System (INIS)

    Bardos, R.; Gorfine, G.; Guy, L.; Moorhead, G.; Taylor, G.; Tovey, S.

    1992-01-01

    This contribution reports new data on the damage of silicon detectors by low energy (1 MeV) neutrons. The data were taken at the end of 1991. Three exposures of UA2 Inner Silicon detectors were made: at +20 deg C, -15 deg C and -95 deg C. A high neutron flux enabled the required fluences to be achieved in relatively short times. This increases the sensitivity of the experiment to damage types with shorter self-annealing time constants. This note discusses the new data taken at +20 deg C. Analysis of the low temperature exposures is in progress. 5 refs., 15 figs

  13. Metal artifact reduction for flat panel detector intravenous CT angiography in patients with intracranial metallic implants after endovascular and surgical treatment.

    Science.gov (United States)

    Pjontek, Rastislav; Önenköprülü, Belgin; Scholz, Bernhard; Kyriakou, Yiannis; Schubert, Gerrit A; Nikoubashman, Omid; Othman, Ahmed; Wiesmann, Martin; Brockmann, Marc A

    2016-08-01

    Flat panel detector CT angiography with intravenous contrast agent injection (IV CTA) allows high-resolution imaging of cerebrovascular structures. Artifacts caused by metallic implants like platinum coils or clips lead to degradation of image quality and are a significant problem. To evaluate the influence of a prototype metal artifact reduction (MAR) algorithm on image quality in patients with intracranial metallic implants. Flat panel detector CT after intravenous application of 80 mL contrast agent was performed with an angiography system (Artis zee; Siemens, Forchheim, Germany) using a 20 s rotation protocol (200° rotation angle, 20 s acquisition time, 496 projections). The data before and after MAR of 26 patients with a total of 34 implants (coils, clips, stents) were independently evaluated by two blinded neuroradiologists. MAR improved the assessability of the brain parenchyma and small vessels (diameter metallic implants and at a distance of 6 cm (p<0.001 each, Wilcoxon test). Furthermore, MAR significantly improved the assessability of parent vessel patency and potential aneurysm remnants (p<0.005 each, McNemar test). MAR, however, did not improve assessability of stented vessels. When an intravenous contrast protocol is used, MAR significantly ameliorates the assessability of brain parenchyma, vessels, and treated aneurysms in patients with intracranial coils or clips. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

  14. Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model

    International Nuclear Information System (INIS)

    Struffert, Tobias; Ott, Sabine; Adamek, Edyta; Schwarz, Marc; Engelhorn, Tobias; Kloska, Stephan; Doerfler, Arnd; Deuerling-Zheng, Yu

    2011-01-01

    Careful follow up is necessary after intracranial stenting because in-stent restenosis (ISR) or residual stenosis (RS) is not rare. A minimally invasive follow-up imaging technique is desirable. The objective was to compare the visualisation of stents in Flat Detector-CT Angiography (FD-CTA) after intravenous contrast medium injection (i.v.) with Multi Detector Computed Tomography Angiography (MD-CTA) and Digital Subtracted Angiography (DSA) in an animal model. Stents were implanted in the carotid artery of 12 rabbits. In 6 a residual stenosis (RS) was surgically created. Imaging was performed using FD-CTA, MD-CTA and DSA. Measurements of the inner and outer diameter and cross-section area of the stents were performed. Stenosis grade was calculated. In subjective evaluation FD-CTA was superior to MD-CTA. FD-CTA was more accurate compared with DSA than MD-CTA. Cross-sectional area of the stent lumen was significantly larger (p < 0.05) in FD-CTA in comparison to MD-CTA. Accurate evaluation of stenosis was impossible in MD-CTA. There was no statistically significant difference in the stenosis grade of DSA and FD-CTA. Our results show that visualisation of stent and stenosis using intravenous FD-CTA compares favourably with DSA and may replace DSA in the follow-up of patients treated with intracranial stents. (orig.)

  15. New detectors technology for radiology imaging

    International Nuclear Information System (INIS)

    Cuzin, M.; Peyret, O.

    1998-01-01

    We summarize the main parameters which describes the radiological image at first and the advantages of pixel detectors. All detectors converts X-rays in charges either with an intermediate step with light or directly in a semi-conductor media. That is true for tomography which is the first domain where digital processing have been taken in account and for radiology where flat panel are now proposed to radiologists. Nevertheless, luminescent stimulated screens are a good way to prepare users with digital radiography. As such technique is not valuable for dynamic acquisition, we describe systems which used standard luminescent screens with CCD cameras or with IIR. Some description and comparison of flat panel independent pixel detectors are given. (authors)

  16. Relations between Biomechanical Parameters and Static Power of Arms in Children with Disturbed Posture

    Directory of Open Access Journals (Sweden)

    Slobodan Andrašić

    2017-02-01

    Full Text Available This study is aimed at determining the parameters and biomechanical analysis of their impact on the static arm strength in children with impaired posture as poor kyphotic posture, lordotic poor posture and children with flat feet. A transversal study included a sample of 67 children on the territory of the municipality of Subotica. The structure of the sample is as follows: 22 subjects with impaired kyphotic posture, 18 patients with impaired lordotic posture, and 27 subjects with flat feet. Measuring the level of static arm strength was done by the standardized "folding endurance" test. Observing the morphological development of children with kyphotic, lordotic poor posture and flat feet determined statistically significant differences in biomechanical variables.

  17. Task-driven orbit design and implementation on a robotic C-arm system for cone-beam CT

    Science.gov (United States)

    Ouadah, S.; Jacobson, M.; Stayman, J. W.; Ehtiati, T.; Weiss, C.; Siewerdsen, J. H.

    2017-03-01

    Purpose: This work applies task-driven optimization to the design of non-circular orbits that maximize imaging performance for a particular imaging task. First implementation of task-driven imaging on a clinical robotic C-arm system is demonstrated, and a framework for orbit calculation is described and evaluated. Methods: We implemented a task-driven imaging framework to optimize orbit parameters that maximize detectability index d'. This framework utilizes a specified Fourier domain task function and an analytical model for system spatial resolution and noise. Two experiments were conducted to test the framework. First, a simple task was considered consisting of frequencies lying entirely on the fz-axis (e.g., discrimination of structures oriented parallel to the central axial plane), and a "circle + arc" orbit was incorporated into the framework as a means to improve sampling of these frequencies, and thereby increase task-based detectability. The orbit was implemented on a robotic C-arm (Artis Zeego, Siemens Healthcare). A second task considered visualization of a cochlear implant simulated within a head phantom, with spatial frequency response emphasizing high-frequency content in the (fy, fz) plane of the cochlea. An optimal orbit was computed using the task-driven framework, and the resulting image was compared to that for a circular orbit. Results: For the fz-axis task, the circle + arc orbit was shown to increase d' by a factor of 1.20, with an improvement of 0.71 mm in a 3D edge-spread measurement for edges located far from the central plane and a decrease in streak artifacts compared to a circular orbit. For the cochlear implant task, the resulting orbit favored complementary views of high tilt angles in a 360° orbit, and d' was increased by a factor of 1.83. Conclusions: This work shows that a prospective definition of imaging task can be used to optimize source-detector orbit and improve imaging performance. The method was implemented for execution of

  18. Seismic restraint means for radiation detector

    International Nuclear Information System (INIS)

    Underwood, R.H.; Todt, W.H.

    1983-01-01

    Seismic restraint means are provided for mounting an elongated, generally cylindrical nuclear radiation detector within a tubular thimble in a nuclear reactor monitor system. The restraint means permits longitudinal movement of the radiation detector into and out of the thimble. Each restraint means comprises a split clamp ring and a plurality of symmetrically spaced support arms pivotally mounted on the clamp ring. Each support arm has spring bias means and thimble contact means eg insulating rollers whereby the contact means engage the thimble with a constant predetermined force which minimizes seismic vibration action on the radiation detector. (author)

  19. Flat detector computed tomography in diagnostic and interventional pediatric cardiology

    International Nuclear Information System (INIS)

    Moesler, J.; Dittrich, S.; Gloeckler, M.; Rompel, O.

    2013-01-01

    Purpose: In this study the use of flat detector computed tomography (FD-CT) in the catheterization of patients with congenital heart disease was evaluated. Application reports were created for various issues based on the achieved image quality in diverse anatomical regions. Materials and Methods: FD-CT was applied in 176 cases during catheterization between January 2010 and April 2012. A five-point Likert scale ('essential' to 'misleading') was used to evaluate image quality. All cases were analyzed retrospectively and application reports for the visualization of the aorta, pulmonary arteries, pulmonary veins, semilunar valves, cavopulmonary connections and atrial baffles were generated. Contrast dye consumption and radiation dose were evaluated. Results: During the observation period FD-CT was applied in all 176 cases. The mean patient age was 7.0 years (0.01 - 42.53 years). The clinical value of FD-CT was rated superior to conventional angiography in 96.6 % of the cases and was never rated as 'misleading'. FD-CT was rated 'essential' in 3.4 % of all cases, 'very useful' in 77.3 % of all cases, 'useful' in 15.9 % of all cases and 'not useful' in 3.4 % of all cases. The mean dose-area product was 99 μGym 2 (19.3 - 1276.6 μGym 2 ), and the used contrast dye was 1.76 ml/kg (0.9 - 5 ml/kg). Application reports for the visualization of different anatomical regions are demonstrated. Conclusion: FD-CT is a new and auxiliary procedure in diagnostic and interventional catheterization of patients with congenital heart disease. Particularly extracardiac structures can be displayed in three-dimensional high resolution and be used for diagnosis, surgical planning and 3 D navigation. (orig.)

  20. Measuring and modeling exposure from environmental radiation on tidal flats

    International Nuclear Information System (INIS)

    Gould, T.J.; Hess, C.T.

    2005-01-01

    To examine the shielding effects of the tide cycle, a high pressure ion chamber was used to measure the exposure rate from environmental radiation on tidal flats. A theoretical model is derived to predict the behavior of exposure rate as a function of time for a detector placed one meter above ground on a tidal flat. The numerical integration involved in this derivation results in an empirical formula which implies exposure rate ∝tan-1(sint). We propose that calculating the total exposure incurred on a tidal flat requires measurements of only the slope of the tidal flat and the exposure rate when no shielding occurs. Experimental results are consistent with the model

  1. Informatics in radiology: use of a C-arm fluoroscopy simulator to support training in intraoperative radiography.

    Science.gov (United States)

    Bott, Oliver Johannes; Dresing, Klaus; Wagner, Markus; Raab, Björn-Werner; Teistler, Michael

    2011-01-01

    Mobile image intensifier systems (C-arms) are used frequently in orthopedic and reconstructive surgery, especially in trauma and emergency settings, but image quality and radiation exposure levels may vary widely, depending on the extent of the C-arm operator's knowledge and experience. Current training programs consist mainly of theoretical instruction in C-arm operation, the physical foundations of radiography, and radiation avoidance, and are largely lacking in hands-on application. A computer-based simulation program such as that tested by the authors may be one way to improve the effectiveness of C-arm training. In computer simulations of various scenarios commonly encountered in the operating room, trainees using the virtX program interact with three-dimensional models to test their knowledge base and improve their skill levels. Radiographs showing the simulated patient anatomy and surgical implants are "reconstructed" from data computed on the basis of the trainee's positioning of models of a C-arm, patient, and table, and are displayed in real time on the desktop monitor. Trainee performance is signaled in real time by color graphics in several control panels and, on completion of the exercise, is compared in detail with the performance of an expert operator. Testing of this computer-based training program in continuing medical education courses for operating room personnel showed an improvement in the overall understanding of underlying principles of intraoperative radiography performed with a C-arm, with resultant higher image quality, lower overall radiation exposure, and greater time efficiency. Supplemental material available at http://radiographics.rsna.org/lookup/suppl/doi:10.1148/rg.313105125/-/DC1. Copyright © RSNA, 2011.

  2. The rotate-plus-shift C-arm trajectory. Part I. Complete data with less than 180° rotation

    International Nuclear Information System (INIS)

    Ritschl, Ludwig; Fleischmann, Christof; Kuntz, Jan; Kachelrieß, Marc

    2016-01-01

    Purpose: In the last decade, C-arm-based cone-beam CT became a widely used modality for intraoperative imaging. Typically a C-arm CT scan is performed using a circular or elliptical trajectory around a region of interest. Therefore, an angular range of at least 180° plus fan angle must be covered to ensure a completely sampled data set. However, mobile C-arms designed with a focus on classical 2D applications like fluoroscopy may be limited to a mechanical rotation range of less than 180° to improve handling and usability. The method proposed in this paper allows for the acquisition of a fully sampled data set with a system limited to a mechanical rotation range of at least 180° minus fan angle using a new trajectory design. This enables CT like 3D imaging with a wide range of C-arm devices which are mainly designed for 2D imaging. Methods: The proposed trajectory extends the mechanical rotation range of the C-arm system with two additional linear shifts. Due to the divergent character of the fan-beam geometry, these two shifts lead to an additional angular range of half of the fan angle. Combining one shift at the beginning of the scan followed by a rotation and a second shift, the resulting rotate-plus-shift trajectory enables the acquisition of a completely sampled data set using only 180° minus fan angle of rotation. The shifts can be performed using, e.g., the two orthogonal positioning axes of a fully motorized C-arm system. The trajectory was evaluated in phantom and cadaver examinations using two prototype C-arm systems. Results: The proposed trajectory leads to reconstructions without limited angle artifacts. Compared to the limited angle reconstructions of 180° minus fan angle, image quality increased dramatically. Details in the rotate-plus-shift reconstructions were clearly depicted, whereas they are dominated by artifacts in the limited angle scan. Conclusions: The method proposed here employs 3D imaging using C-arms with less than 180° rotation

  3. The rotate-plus-shift C-arm trajectory. Part I. Complete data with less than 180° rotation

    Energy Technology Data Exchange (ETDEWEB)

    Ritschl, Ludwig; Fleischmann, Christof [Ziehm Imaging GmbH, Donaustraße 31, Nürnberg 90451 (Germany); Kuntz, Jan, E-mail: j.kuntz@dkfz.de; Kachelrieß, Marc [Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120 (Germany)

    2016-05-15

    Purpose: In the last decade, C-arm-based cone-beam CT became a widely used modality for intraoperative imaging. Typically a C-arm CT scan is performed using a circular or elliptical trajectory around a region of interest. Therefore, an angular range of at least 180° plus fan angle must be covered to ensure a completely sampled data set. However, mobile C-arms designed with a focus on classical 2D applications like fluoroscopy may be limited to a mechanical rotation range of less than 180° to improve handling and usability. The method proposed in this paper allows for the acquisition of a fully sampled data set with a system limited to a mechanical rotation range of at least 180° minus fan angle using a new trajectory design. This enables CT like 3D imaging with a wide range of C-arm devices which are mainly designed for 2D imaging. Methods: The proposed trajectory extends the mechanical rotation range of the C-arm system with two additional linear shifts. Due to the divergent character of the fan-beam geometry, these two shifts lead to an additional angular range of half of the fan angle. Combining one shift at the beginning of the scan followed by a rotation and a second shift, the resulting rotate-plus-shift trajectory enables the acquisition of a completely sampled data set using only 180° minus fan angle of rotation. The shifts can be performed using, e.g., the two orthogonal positioning axes of a fully motorized C-arm system. The trajectory was evaluated in phantom and cadaver examinations using two prototype C-arm systems. Results: The proposed trajectory leads to reconstructions without limited angle artifacts. Compared to the limited angle reconstructions of 180° minus fan angle, image quality increased dramatically. Details in the rotate-plus-shift reconstructions were clearly depicted, whereas they are dominated by artifacts in the limited angle scan. Conclusions: The method proposed here employs 3D imaging using C-arms with less than 180° rotation

  4. Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg

    Energy Technology Data Exchange (ETDEWEB)

    Miraglia, Roberto; Maruzzelli, Luigi; Cortis, Kelvin; Gerasia, Roberta; Maggio, Simona; Luca, Angelo [Diagnostic and Therapeutic Services, Mediterranean Institute for Transplantation and Advanced Specialized Therapies (ISMETT), Palermo (Italy); Piazza, Marcello [Department of Anesthesia, Mediterranean Institute for Transplantation and Advanced Specialized Therapies (ISMETT), Palermo (Italy); Tuzzolino, Fabio [Department of Information Technology, Mediterranean Institute for Transplantation and Advanced Specialized Therapies (ISMETT), Palermo (Italy)

    2014-09-10

    Ultrasound-guided central venous puncture and fluoroscopic guidance during central venous catheter (CVC) positioning optimizes technical success and lowers the complication rates in children, and is therefore considered standard practice. The purpose of this study was to compare the radiation exposure levels recorded during CVC placement in children weighing less than 10 kg in procedures performed using an image intensifier-based angiographic system (IIDS) to those performed in a flat-panel detector-based interventional suite (FPDS). A retrospective review of 96 image-guided CVC placements, between January 2008 and October 2013, in 49 children weighing less than 10 kg was performed. Mean age was 8.2 ± 4.4 months (range: 1-22 months). Mean weight was 7.1 ± 2.7 kg (range: 2.5-9.8 kg). The procedures were classified into two categories: non-tunneled and tunneled CVC placement. Thirty-five procedures were performed with the IIDS (21 non-tunneled CVC, 14 tunneled CVC); 61 procedures were performed with the FPDS (47 non-tunneled CVC, 14 tunneled CVC). For non-tunneled CVC, mean DAP was 113.5 ± 126.7 cGy cm{sup 2} with the IIDS and 15.9 ± 44.6 cGy . cm{sup 2} with the FPDS (P < 0.001). For tunneled CVC, mean DAP was 84.6 ± 81.2 cGy . cm{sup 2} with the IIDS and 37.1 ± 33.5 cGy cm{sup 2} with the FPDS (P = 0.02). The use of flat-panel angiographic equipment reduces radiation exposure in small children undergoing image-guided CVC placement. (orig.)

  5. Characterization of a SiC MIS Schottky diode as RBS particle detector

    Science.gov (United States)

    Kaufmann, I. R.; Pick, A. C.; Pereira, M. B.; Boudinov, H. I.

    2018-02-01

    A 4H-SiC Schottky diode was investigated as a particle detector for Rutherford Backscattering Spectroscopy (RBS) experiment. The device was fabricated on a commercial 4H-SiC epitaxial n-type layer grown onto a 4H-SiC n+ type substrate wafer doped with nitrogen. Hafnium oxide with thickness of 1 nm was deposited by Atomic Layer Deposition and 10 nm of Ni were deposited by sputtering to form the Ni/HfO2/4H-SiC MIS Schottky structure. Current-Voltage curves with variable temperature were measured to extract the real Schottky Barrier Height (0.32 V) and ideality factor values (1.15). Reverse current and Capacitance-Voltage measurements were performed on the 4H-SiC detector and compared to a commercial Si barrier detector acquired from ORTEC. RBS data for four alpha energies (1, 1.5, 2 and 2.5 MeV) were collected from an Au/Si sample using the fabricated SiC and the commercial Si detectors simultaneously. The energy resolution for the fabricated detector was estimated to be between 75 and 80 keV.

  6. Study on the method or reducing the operator's exposure dose from a C-Arm system

    International Nuclear Information System (INIS)

    Kim, Ki Sik; Song, Jong Nam; Kim, Seung Ok

    2016-01-01

    In this study, C-Arm equipment is being used as we intend to verify the exposure dose on the operator by the scattering rays during the operation of the C-Arm equipment and to provide an effective method of reducing the exposure dose. Exposure dose is less than the Over Tube method utilizes the C-arm equipment Under Tube the scheme, The result showed that the exposure dose on the operator decreased with a thicker shield, and as the operator moved away from the center line. Moreover, as the research time prolongated, the exposure dose increased, and among the three affixed location of the dosimeter, the most exposure dose was measured at gonadal, then followed by chest and thyroid. However, in consideration of the relationship between the operator and the patient, the distance cannot be increased infinitely and the research time cannot be decreased infinitely in order to reduce the exposure dose. Therefore, by changing the thickness of the radiation shield, the exposure dose on the operator was able to be reduced. If you are using a C-Arm equipment discomfort during surgery because the grounds that the procedure is neglected and close to the dose of radiation shielding made can only increase. Because a separate control room cannot be used for the C-Arm equipment due to its characteristic, the exposure dose on the operator needs to be reduced by reinforcing the shield through an appropriate thickness of radiation shield devices, such as apron, etc. during a treatment

  7. Dose-equivalent response CR-39 track detector for personnel neutron dosimetry

    International Nuclear Information System (INIS)

    Oda, K.; Ito, M.; Yoneda, H.; Miyake, H.; Yamamoto, J.; Tsuruta, T.

    1991-01-01

    A dose-equivalent response detector based on CR-39 has been designed to be applied for personnel neutron dosimetry. The intrinsic detection efficiency of bare CR-39 was first evaluated from irradiation experiments with monoenergetic neutrons and theoretical calculations. In the second step, the radiator effect was investigated for the purpose of sensitization to fast neutrons. A two-layer radiator consisting of deuterized dotriacontane (C 32 D 66 ) and polyethylene (CH 2 ) was designed. Finally, we made the CR-39 detector sensitive to thermal neutrons by doping with orthocarbone (B 10 H 12 C 2 ), and also estimated the contribution of albedo neutrons. It was found that the new detector - boron-doped CR-39 with the two-layer radiator - would have a flat response with an error of about 70% in a wide energy region, ranging from thermal to 15 MeV. (orig.)

  8. Camera-augmented mobile C-arm (CamC): A feasibility study of augmented reality imaging in the operating room.

    Science.gov (United States)

    von der Heide, Anna Maria; Fallavollita, Pascal; Wang, Lejing; Sandner, Philipp; Navab, Nassir; Weidert, Simon; Euler, Ekkehard

    2018-04-01

    In orthopaedic trauma surgery, image-guided procedures are mostly based on fluoroscopy. The reduction of radiation exposure is an important goal. The purpose of this work was to investigate the impact of a camera-augmented mobile C-arm (CamC) on radiation exposure and the surgical workflow during a first clinical trial. Applying a workflow-oriented approach, 10 general workflow steps were defined to compare the CamC to traditional C-arms. The surgeries included were arbitrarily identified and assigned to the study. The evaluation criteria were radiation exposure and operation time for each workflow step and the entire surgery. The evaluation protocol was designed and conducted in a single-centre study. The radiation exposure was remarkably reduced by 18 X-ray shots 46% using the CamC while keeping similar surgery times. The intuitiveness of the system, its easy integration into the surgical workflow, and its great potential to reduce radiation have been demonstrated. Copyright © 2017 John Wiley & Sons, Ltd.

  9. The value of flat-detector computed tomography during catheterisation of congenital heart disease

    International Nuclear Information System (INIS)

    Gloeckler, Martin; Koch, Andreas; Greim, Verena; Shabaiek, Amira; Dittrich, Sven; Rueffer, Andre; Cesnjevar, Robert; Achenbach, Stephan

    2011-01-01

    To analyse the diagnostic utility of flat-detector computed tomography imaging (FD-CT) in patients with congenital heart disease, including the value of image fusion to overlay three-dimensional (3D) reconstructions on fluoroscopic images during catheter-based interventions. We retrospectively analysed 62 consecutive paediatric patients in whom FD-CT was used during catheterisation of congenital heart disease. Expert operators rated the clinical value of FD-CT over conventional fluoroscopic imaging. Added radiation exposure and contrast medium volume were evaluated. During a 12-month period, FD-CT was performed in 62 out of 303 cardiac catheterisations. Median patient age was 3.5 years. In 32/62 cases, FD-CT was used for diagnostic purposes, in 30/62 cases it was used in the context of interventions. Diagnostic utility was never rated as ''misleading''. It was classified as ''not useful'' in six cases (9.7%), ''useful'' in 18 cases (29.0%), ''very useful'' in 37 cases (59.7%) and ''essential'' in one case (1.6%). The median added dose-area product was 111.0 μGym 2 , the required additional quantity of contrast medium was 1.6 ml/kg. FD-CT provides useful diagnostic information in most of the patients investigated for congenital heart disease. The added radiation exposure and contrast medium volume are reasonable. (orig.)

  10. Application of C-arm computed tomography in cardiology

    International Nuclear Information System (INIS)

    Rieber, J.; Rohkohl, C.; Lauritsch, G.; Rittger, H.; Meissner, O.

    2009-01-01

    C-arm computed tomography is currently being introduced into cardiac imaging and offers the potential for three-dimensional imaging of the cardiac anatomy within the interventional environment. This detailed view is necessary to support complex interventional strategies, such as transcutaneous valve replacement, interventional therapy of atrial fibrillation, implantation of biventricular pacemakers and assessment of myocardial perfusion. Currently, the major limitation of this technology is its insufficient temporal resolution which limits the visualization of fast moving parts of the heart. (orig.) [de

  11. The nuclear data collecting system designed with ARM and μC/OS-II

    International Nuclear Information System (INIS)

    Wang Chunsheng; Ma Yingjie; Han Feng

    2008-01-01

    Introduce a kind of nuclear data collecting system regarding ARM-μC/OS-II as the platform, gathering the GPS receiver in it. It was detailed to expatiated the key techniques of the multi-channel pulse amplitude analyzer, the interface design of LPC2148, a controller in ARM, and how to expand the RTOS and design applications on μC/OS-II. This system can communicate with the GPS-OEM module by the UART interface, collecting the GPS information synchronously as well as nuclear data. And then save and display them or pass them to the host computer by the USB interface. The embedded and Real-Time system, μC/OS-II build up the real-time and stability of the system and advance the integration. (authors)

  12. 14C autoradiography with an energy-sensitive silicon pixel detector.

    Science.gov (United States)

    Esposito, M; Mettivier, G; Russo, P

    2011-04-07

    The first performance tests are presented of a carbon-14 ((14)C) beta-particle digital autoradiography system with an energy-sensitive hybrid silicon pixel detector based on the Timepix readout circuit. Timepix was developed by the Medipix2 Collaboration and it is similar to the photon-counting Medipix2 circuit, except for an added time-based synchronization logic which allows derivation of energy information from the time-over-threshold signal. This feature permits direct energy measurements in each pixel of the detector array. Timepix is bump-bonded to a 300 µm thick silicon detector with 256 × 256 pixels of 55 µm pitch. Since an energetic beta-particle could release its kinetic energy in more than one detector pixel as it slows down in the semiconductor detector, an off-line image analysis procedure was adopted in which the single-particle cluster of hit pixels is recognized; its total energy is calculated and the position of interaction on the detector surface is attributed to the centre of the charge cluster. Measurements reported are detector sensitivity, (4.11 ± 0.03) × 10(-3) cps mm(-2) kBq(-1) g, background level, (3.59 ± 0.01) × 10(-5) cps mm(-2), and minimum detectable activity, 0.0077 Bq. The spatial resolution is 76.9 µm full-width at half-maximum. These figures are compared with several digital imaging detectors for (14)C beta-particle digital autoradiography.

  13. Noise variance analysis using a flat panel x-ray detector: A method for additive noise assessment with application to breast CT applications

    Energy Technology Data Exchange (ETDEWEB)

    Yang Kai; Huang, Shih-Ying; Packard, Nathan J.; Boone, John M. [Department of Radiology, University of California, Davis Medical Center, 4860 Y Street, Suite 3100 Ellison Building, Sacramento, California 95817 (United States); Department of Radiology, University of California, Davis Medical Center, 4860 Y Street, Suite 3100 Ellison Building, Sacramento, California 95817 (United States) and Department of Biomedical Engineering, University of California, Davis, Davis, California, 95616 (United States)

    2010-07-15

    Purpose: A simplified linear model approach was proposed to accurately model the response of a flat panel detector used for breast CT (bCT). Methods: Individual detector pixel mean and variance were measured from bCT projection images acquired both in air and with a polyethylene cylinder, with the detector operating in both fixed low gain and dynamic gain mode. Once the coefficients of the linear model are determined, the fractional additive noise can be used as a quantitative metric to evaluate the system's efficiency in utilizing x-ray photons, including the performance of different gain modes of the detector. Results: Fractional additive noise increases as the object thickness increases or as the radiation dose to the detector decreases. For bCT scan techniques on the UC Davis prototype scanner (80 kVp, 500 views total, 30 frames/s), in the low gain mode, additive noise contributes 21% of the total pixel noise variance for a 10 cm object and 44% for a 17 cm object. With the dynamic gain mode, additive noise only represents approximately 2.6% of the total pixel noise variance for a 10 cm object and 7.3% for a 17 cm object. Conclusions: The existence of the signal-independent additive noise is the primary cause for a quadratic relationship between bCT noise variance and the inverse of radiation dose at the detector. With the knowledge of the additive noise contribution to experimentally acquired images, system modifications can be made to reduce the impact of additive noise and improve the quantum noise efficiency of the bCT system.

  14. Flat panel computed tomography of human ex vivo heart and bone specimens: initial experience

    Energy Technology Data Exchange (ETDEWEB)

    Nikolaou, Konstantin; Becker, Christoph R.; Reiser, Maximilian F. [Ludwig-Maximilians-University, Department of Clinical Radiology, Munich (Germany); Flohr, Thomas; Stierstorfer, Karl [CT Division, Siemens Medical Solutions, Forchheim (Germany)

    2005-02-01

    The aim of this technical investigation was the detailed description of a prototype flat panel detector computed tomography system (FPCT) and its initial evaluation in an ex vivo setting. The prototype FPCT scanner consists of a conventional radiographic flat panel detector, mounted on a multi-slice CT scanner gantry. Explanted human ex vivo heart and foot specimens were examined. Images were reformatted with various reconstruction algorithms and were evaluated for high-resolution anatomic information. For comparison purposes, the ex vivo specimens were also scanned with a conventional 16-detector-row CT scanner (Sensation 16, Siemens Medical Solutions, Forchheim, Germany). With the FPCT prototype used, a 1,024 x 768 resolution matrix can be obtained, resulting in an isotropic voxel size of 0.25 x 0.25 x 0.25 mm at the iso-center. Due to the high spatial resolution, very small structures such as trabecular bone or third-degree, distal branches of coronary arteries could be visualized. This first evaluation showed that flat panel detector systems can be used in a cone-beam computed tomography scanner and that very high spatial resolutions can be achieved. However, there are limitations for in vivo use due to constraints in low contrast resolution and slow scan speed. (orig.)

  15. 2D-3D radiograph to cone-beam computed tomography (CBCT) registration for C-arm image-guided robotic surgery.

    Science.gov (United States)

    Liu, Wen Pei; Otake, Yoshito; Azizian, Mahdi; Wagner, Oliver J; Sorger, Jonathan M; Armand, Mehran; Taylor, Russell H

    2015-08-01

    C-arm radiographs are commonly used for intraoperative image guidance in surgical interventions. Fluoroscopy is a cost-effective real-time modality, although image quality can vary greatly depending on the target anatomy. Cone-beam computed tomography (CBCT) scans are sometimes available, so 2D-3D registration is needed for intra-procedural guidance. C-arm radiographs were registered to CBCT scans and used for 3D localization of peritumor fiducials during a minimally invasive thoracic intervention with a da Vinci Si robot. Intensity-based 2D-3D registration of intraoperative radiographs to CBCT was performed. The feasible range of X-ray projections achievable by a C-arm positioned around a da Vinci Si surgical robot, configured for robotic wedge resection, was determined using phantom models. Experiments were conducted on synthetic phantoms and animals imaged with an OEC 9600 and a Siemens Artis zeego, representing the spectrum of different C-arm systems currently available for clinical use. The image guidance workflow was feasible using either an optically tracked OEC 9600 or a Siemens Artis zeego C-arm, resulting in an angular difference of Δθ:∼ 30°. The two C-arm systems provided TRE mean ≤ 2.5 mm and TRE mean ≤ 2.0 mm, respectively (i.e., comparable to standard clinical intraoperative navigation systems). C-arm 3D localization from dual 2D-3D registered radiographs was feasible and applicable for intraoperative image guidance during da Vinci robotic thoracic interventions using the proposed workflow. Tissue deformation and in vivo experiments are required before clinical evaluation of this system.

  16. Neutron-sensitive ZnS/10B2O3 ceramic scintillator detector as an alternative to a 3He-gas-based detector for a plutonium canister assay system

    International Nuclear Information System (INIS)

    Nakamura, T.; Ohzu, A.; Toh, K.; Sakasai, K.; Suzuki, H.; Honda, K.; Birumachi, A.; Ebine, M.; Yamagishi, H.; Takase, M.; Haruyama, M.; Kureta, M.; Soyama, K.; Nakamura, H.; Seya, M.

    2014-01-01

    A neutron-sensitive ZnS/ 10 B 2 O 3 ceramic scintillator detector was developed as an alternative to a 3 He-gas-based detector for use in a plutonium canister assay system. The detector has a modular structure, with a flat ZnS/ 10 B 2 O 3 ceramic scintillator strip that is installed diagonally inside a light-reflecting aluminium case with a square cross-section, and where the scintillation light is detected using two photomultiplier tubes attached at both ends of the case. The prototype detectors, which have a neutron-sensitive area of 30 mm×250 mm, exhibited a sensitivity of 21.7–23.4±0.1 cps/nv (mean±SD) for thermal neutrons, a 137 Cs gamma-ray sensitivity of 1.1–1.9±0.2×10 −7 and a count variation of less than 6% over the detector length. A trial experiment revealed a temperature coefficient of less than −0.24±0.05%/°C over the temperature range of 20–50 °C. The detector design and the experimental results are presented

  17. Tests of innovative photon detectors and integrated electronics for the large-area CLAS12 ring-imaging Cherenkov detector

    Energy Technology Data Exchange (ETDEWEB)

    Contalbrigo, M., E-mail: contalbrigo@fe.infn.it

    2015-07-01

    A large area ring-imaging Cherenkov detector has been designed to provide clean hadron identification capability in the momentum range from 3 GeV/c to 8 GeV/c for the CLAS12 experiments at the upgraded 12 GeV continuous electron beam accelerator facility of Jefferson Lab. Its aim is to study the 3D nucleon structure in the yet poorly explored valence region by deep-inelastic scattering, and to perform precision measurements in hadron spectroscopy. The adopted solution foresees a novel hybrid optics design based on an aerogel radiator, composite mirrors and a densely packed and highly segmented photon detector. Cherenkov light will either be imaged directly (forward tracks) or after two mirror reflections (large angle tracks). Extensive tests have been performed on Hamamatsu H8500 and novel flat multi-anode photomultipliers under development and on various types of silicon photomultipliers. A large scale prototype based on 28 H8500 MA-PMTs has been realized and tested with few GeV/c hadron beams at the T9 test-beam facility of CERN. In addition a small prototype was used to study the response of customized SiPM matrices within a temperature interval ranging from 25 down to −25 °C. The preliminary results of the individual photon detector tests and of the prototype performance at the test-beams are here reported.

  18. Radioactive flow detectors: liquid or solid scintillators

    International Nuclear Information System (INIS)

    Reich, A.R.

    1983-01-01

    During the past five years, two schools of thought have emerged producing two different types of radio-HPLC detectors. Based on the naphthalene-in-the-vial principle, manufacturers have developed heterogeneous scintillation detectors. In these detectors the anthracene or naphthalene crystals are replaced by other scintillators. In order to avoid dead space and turbulence, a narrow diameter tube is used, either straight, or more popularly formed into a coil or a 'U' as the cell. To optimize light transmission to the photomultiplier tubes, mirrors are used. Due to limiting factors in this technique the counting efficiency for tritium is below the 10 percent level. The other school of radio-HPLC detectors based their design on classical liquid scintillation counting technology. In a homogeneous detector, the effluent from the HPLC system is mixed with a suitable liquid scintillator before entering the counting cell. The cell design is typically a flat glass or Teflon coil tightly sandwiched between two photomultiplier tubes, making good optical contact without the use of mirrors. Depending on the chromatographic effluent, 3 H efficiencies between 25 to 50 percent, and 14 C counting efficiencies up to 85 percent can be achieved

  19. Properties of SiC semiconductor detector of fast neutrons investigated using MCNPX code

    International Nuclear Information System (INIS)

    Sedlakova, K.; Sagatova, A.; Necas, V.; Zatko, B.

    2013-01-01

    The potential of silicon carbide (SiC) for use in semiconductor nuclear radiation detectors has been long recognized. The wide bandgap of SiC (3.25 eV for 4H-SiC polytype) compared to that for more conventionally used semiconductors, such as silicon (1.12 eV) and germanium (0.67 eV), makes SiC an attractive semiconductor for use in high dose rate and high ionization nuclear environments. The present work focused on the simulation of particle transport in SiC detectors of fast neutrons using statistical analysis of Monte Carlo radiation transport code MCNPX. Its possibilities in detector design and optimization are presented.(authors)

  20. Development of SiC Neutron Detector Assembly to Measure the Neutron Flux of the Reactor Core

    Energy Technology Data Exchange (ETDEWEB)

    Park, Se Hwan; Park, June Sic; Shin, Hee Sung; Kim, Ho Dong [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of); Kim, Yong Kyun [Hanyang University, Seoul (Korea, Republic of)

    2012-05-15

    At present, the conventional detector to measure the neutron at harsh environment is a Self Powered Neutron Detector (SPND). Rhodium(Rh)-103 is in the SPND. When neutron is incident on the Rhodium, the neutron capture reaction occurs, and the Rh-103 is converted to Rh-104. The Rh-104 is decayed to Pd-104 by {beta}-decay, and electrons are generated as the decay products. Because of the half life of Rh-104, approximately 5 minutes are required for the SPND output to reach the equilibrium condition. Therefore the on-line monitoring of the nuclear reactor state is limited if the neutron flux in the reactor core is monitored with the SPND. Silicon carbide (SiC) has the possibility to be developed as neutron detector at harsh environment, because the SiC can be operative at high temperature and high neutron flux conditions. Previously, the basic operation properties of the SiC detector were studied. Also, the radiation response of the SiC detector was studied at high neutron and gamma dose rate. The measurement results for an ex-core neutron flux monitor or a neutron flux monitor of the spent fuel were published. The SiC detector was also developed as neutron detector to measure the fissile material with active interrogation method. However, the studies about the development of SiC detector are still limited. In the present work, the radiation damage effect of the SiC detector was studied. The detector structure was determined based on the study, and a neutron detector assembly was made with the SiC detectors. The neutron and gamma-ray response of the detector assembly is presented in this paper. The detector assembly was positioned in the HANARO research reactor core, the performance test was done. The preliminary results are also included in this paper

  1. Comparison of C-arm computed tomography and on-site quick cortisol assay for adrenal venous sampling: A retrospective study of 178 patients

    Energy Technology Data Exchange (ETDEWEB)

    Chang, Chin-Chen; Lee, Bo-Ching; Chang, Yeun-Chung; Liu, Kao-Lang [National Taiwan University Hospital and National Taiwan University College of Medicine, Department of Medical Imaging, Taipei (China); Wu, Vin-Cent [National Taiwan University Hospital and National Taiwan University College of Medicine, Department of Internal Medicine, Taipei (China); Huang, Kuo-How [National Taiwan University Hospital and National Taiwan University College of Medicine, Department of Urology, Taipei (China); Collaboration: on behalf of the TAIPAI Study Group

    2017-12-15

    To compare the performance of on-site quick cortisol assay (QCA) and C-arm computed tomography (CT) assistance on adrenal venous sampling (AVS) without adrenocorticotropic hormone stimulation. The institutional review board at our hospital approved this retrospective study, which included 178 consecutive patients with primary aldosteronism. During AVS, we used C-arm CT to confirm right adrenal cannulation between May 2012 and June 2015 (n = 100) and QCA for bilateral adrenal cannulation between July 2015 and September 2016 (n = 78). Successful AVS required a selectivity index (cortisol{sub adrenal} {sub vein}/cortisol{sub peripheral}) of ≥ 2.0 bilaterally. The overall success rate of C-arm CT-assisted AVS was 87%, which increased to 97.4% under QCA (P =.013). The procedure time (C-arm CT, 49.5 ± 21.3 min; QCA, 37.5 ± 15.6 min; P <.001) and radiation dose (C-arm CT, 673.9 ± 613.8 mGy; QCA, 346.4 ± 387.8 mGy; P <.001) were also improved. The resampling rate was 16% and 21.8% for C-arm CT and QCA, respectively. The initial success rate of the performing radiologist remained stable during the study period (C-arm CT 75%; QCA, 82.1%, P =.259). QCA might be superior to C-arm CT for improving the performance of AVS. (orig.)

  2. A novel method for contrast-to-noise ratio (CNR) evaluation of digital mammography detectors

    International Nuclear Information System (INIS)

    Baldelli, P.; Phelan, N.; Egan, G.

    2009-01-01

    The purpose of this study was to test a new, simple method of evaluating the contrast-to-noise ratio (CNR) over the entire image field of a digital detector and to compare different mammography systems. Images were taken under clinical exposure conditions for a range of simulated breast thicknesses using poly(methyl methacrylate) (PMMA). At each PMMA thickness, a second image which included an additional 0.2-mm Al sheet was also acquired. Image processing software was used to calculate the CNR in multiple regions of interest (ROI) covering the entire area of the detector in order to obtain a 'CNR image'. Five detector types were evaluated, two CsI-αSi (GE Healthcare) flat panel systems, one αSe (Hologic) flat panel system and a two generations of scanning photon counting digital detectors (Sectra). Flat panel detectors exhibit better CNR uniformity compared with the first-generation scanning photon counting detector in terms of mean pixel value variation. However, significant improvement in CNR uniformity was observed for the next-generation scanning detector. The method proposed produces a map of the CNR and a measurement of uniformity throughout the entire image field of the detector. The application of this method enables quality control measurement of individual detectors and a comparison of detectors using different technologies. (orig.)

  3. Large area x-ray detectors for cargo radiography

    Science.gov (United States)

    Bueno, C.; Albagli, D.; Bendahan, J.; Castleberry, D.; Gordon, C.; Hopkins, F.; Ross, W.

    2007-04-01

    Large area x-ray detectors based on phosphors coupled to flat panel amorphous silicon diode technology offer significant advances for cargo radiologic imaging. Flat panel area detectors provide large object coverage offering high throughput inspections to meet the high flow rate of container commerce. These detectors provide excellent spatial resolution when needed, and enhanced SNR through low noise electronics. If the resolution is reduced through pixel binning, further advances in SNR are achievable. Extended exposure imaging and frame averaging enables improved x-ray penetration of ultra-thick objects, or "select-your-own" contrast sensitivity at a rate many times faster than LDAs. The areal coverage of flat panel technology provides inherent volumetric imaging with the appropriate scanning methods. Flat panel area detectors have flexible designs in terms of electronic control, scintillator selection, pixel pitch, and frame rates. Their cost is becoming more competitive as production ramps up for the healthcare, nondestructive testing (NDT), and homeland protection industries. Typically used medical and industrial polycrystalline phosphor materials such as Gd2O2S:Tb (GOS) can be applied to megavolt applications if the phosphor layer is sufficiently thick to enhance x-ray absorption, and if a metal radiator is used to augment the quantum detection efficiency and reduce x-ray scatter. Phosphor layers ranging from 0.2-mm to 1-mm can be "sandwiched" between amorphous silicon flat panel diode arrays and metal radiators. Metal plates consisting of W, Pb or Cu, with thicknesses ranging from 0.25-mm to well over 1-mm can be used by covering the entire area of the phosphor plate. In some combinations of high density metal and phosphor layers, the metal plate provides an intensification of 25% in signal due to electron emission from the plate and subsequent excitation within the phosphor material. This further improves the SNR of the system.

  4. Rotating detectors and Mach's principle

    International Nuclear Information System (INIS)

    Paola, R.D.M. de; Svaiter, N.F.

    2000-08-01

    In this work we consider a quantum version of Newton s bucket experiment in a fl;at spacetime: we take an Unruh-DeWitt detector in interaction with a real massless scalar field. We calculate the detector's excitation rate when it is uniformly rotating around some fixed point and the field is prepared in the Minkowski vacuum and also when the detector is inertial and the field is in the Trocheries-Takeno vacuum state. These results are compared and the relations with Mach's principle are discussed. (author)

  5. Reference handbook: Level detectors

    International Nuclear Information System (INIS)

    1990-01-01

    The purpose of this handbook is to provide Rocky Flats personnel with the information necessary to understand level measurement and detection. Upon completion of this handbook you should be able to do the following: List three reasons for measuring level. Describe the basic operating principles of the sight glass. Demonstrate proper techniques for reading a sight glass. Describe the basic operating principles of a float level detector. Describe the basic operating principles of a bubbler level indicating system. Explain the differences between a wet and dry reference leg indicating system, and describe how each functions. This handbook is designed for use by experienced Rocky Flats operators to reinforce and improve their current knowledge level, and by entry-level operators to ensure that they possess a minimum level of fundamental knowledge. Level Detectors is applicable to many job classifications and can be used as a reference for classroom work or for self-study. Although this reference handbook is by no means all-encompassing, you will gain enough information about this subject area to assist you in contributing to the safe operation of Rocky Flats Plant

  6. Nine-degrees-of-freedom flexmap for a cone-beam computed tomography imaging device with independently movable source and detector.

    Science.gov (United States)

    Keuschnigg, Peter; Kellner, Daniel; Fritscher, Karl; Zechner, Andrea; Mayer, Ulrich; Huber, Philipp; Sedlmayer, Felix; Deutschmann, Heinz; Steininger, Philipp

    2017-01-01

    Couch-mounted cone-beam computed tomography (CBCT) imaging devices with independently rotatable x-ray source and flat-panel detector arms for acquisitions of arbitrary regions of interest (ROI) have recently been introduced in image-guided radiotherapy (IGRT). This work analyzes mechanical limitations and gravity-induced effects influencing the geometric accuracy of images acquired with arbitrary angular constellations of source and detector in nonisocentric trajectories, which is considered essential for IGRT. In order to compensate for geometric inaccuracies of this modality, a 9-degrees-of-freedom (9-DOF) flexmap correction approach is presented, focusing especially on the separability of the flexmap parameters of the independently movable components of the device. The 9-DOF comprise a 3D translation of the x-ray source focal spot, a 3D translation of the flat-panel's active area center and three Euler-rotations of the detector's row and column vectors. The flexmap parameters are expressed with respect to the angular position of each of the devices arms. Estimation of the parameters is performed, using a CT-based structure set of a table-mounted, cylindrical ball-bearing phantom. Digitally reconstructed radiograph (DRR) patches are derived from the structure set followed by local 2D in-plane registration and subsequent 3D transform estimation by nonlinear regression with outlier detection. Flexmap parameter evaluations for the factory-calibrated system in clockwise and counter-clockwise rotation direction have shown only minor differences for the overall set of flexmap parameters. High short-term reproducibility of the flexmap parameters has been confirmed by experiments over 10 acquisitions for both directions, resulting in standard deviation values of ≤0.183 mm for translational components and ≤0.0219 deg for rotational components, respectively. A comparison of isocentric and nonisocentric flexmap evaluations showed that the mean differences of the

  7. The Usefulness of Three-Dimensional Angiography with a Flat Panel Detector of Direct Conversion Type in a Transcatheter Arterial Chemoembolization Procedure for Hepatocellular Carcinoma: Initial Experience

    International Nuclear Information System (INIS)

    Kakeda, Shingo; Korogi, Yukunori; Hatakeyama, Yoshihisa; Ohnari, Norihiro; Oda, Nobuhiro; Nishino, Kazuyoshi; Miyamoto, Wataru

    2008-01-01

    The purpose of this study was to assess the usefulness of a three-dimensional (3D) angiography system using a flat panel detector of direct conversion type in treatments with subsegmental transcatheter arterial chemoembolization (TACE) for hepatocellular carcinomas (HCCs). Thirty-six consecutive patients who underwent hepatic angiography were prospectively examined. First, two radiologists evaluated the degree of visualization of the peripheral branches of the hepatic arteries on 3D digital subtraction angiography (DSA). Then the radiologists evaluated the visualization of tumor staining and feeding arteries in 25 patients (30 HCCs) who underwent subsegmental TACE. The two radiologists who performed the TACE assessed whether the additional information provided by 3D DSA was useful for treatments. In 34 (94.4%) of 36 patients, the subsegmental branches of the hepatic arteries were sufficiently visualized. The feeding arteries of HCCs were sufficiently visualized in 28 (93%) of 30 HCCs, whereas tumor stains were sufficiently visualized in 18 (60%). Maximum intensity projection images were significantly superior to volume recording images for visualization of the tumor staining and feeding arteries of HCCs. In 27 (90%) of 30 HCCs, 3D DSA provided additional useful information for subsegmental TACE. The high-quality 3D DSA with flat panel detector angiography system provided a precise vascular road map, which was useful for performing subsegmental TACE .of HCCs

  8. Study on the method or reducing the operator's exposure dose from a C-Arm system

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Ki Sik; Song, Jong Nam [Dept. of Radiological Science, Dongshin University, Naju (Korea, Republic of); Kim, Seung Ok [Dept. of Radiology, Catholic Kwangdong Universty International ST.Mary' s Hospital, Incheon (Korea, Republic of)

    2016-12-15

    In this study, C-Arm equipment is being used as we intend to verify the exposure dose on the operator by the scattering rays during the operation of the C-Arm equipment and to provide an effective method of reducing the exposure dose. Exposure dose is less than the Over Tube method utilizes the C-arm equipment Under Tube the scheme, The result showed that the exposure dose on the operator decreased with a thicker shield, and as the operator moved away from the center line. Moreover, as the research time prolongated, the exposure dose increased, and among the three affixed location of the dosimeter, the most exposure dose was measured at gonadal, then followed by chest and thyroid. However, in consideration of the relationship between the operator and the patient, the distance cannot be increased infinitely and the research time cannot be decreased infinitely in order to reduce the exposure dose. Therefore, by changing the thickness of the radiation shield, the exposure dose on the operator was able to be reduced. If you are using a C-Arm equipment discomfort during surgery because the grounds that the procedure is neglected and close to the dose of radiation shielding made can only increase. Because a separate control room cannot be used for the C-Arm equipment due to its characteristic, the exposure dose on the operator needs to be reduced by reinforcing the shield through an appropriate thickness of radiation shield devices, such as apron, etc. during a treatment.

  9. Comparison of C-arm computed tomography and on-site quick cortisol assay for adrenal venous sampling: A retrospective study of 178 patients.

    Science.gov (United States)

    Chang, Chin-Chen; Lee, Bo-Ching; Chang, Yeun-Chung; Wu, Vin-Cent; Huang, Kuo-How; Liu, Kao-Lang

    2017-12-01

    To compare the performance of on-site quick cortisol assay (QCA) and C-arm computed tomography (CT) assistance on adrenal venous sampling (AVS) without adrenocorticotropic hormone stimulation. The institutional review board at our hospital approved this retrospective study, which included 178 consecutive patients with primary aldosteronism. During AVS, we used C-arm CT to confirm right adrenal cannulation between May 2012 and June 2015 (n = 100) and QCA for bilateral adrenal cannulation between July 2015 and September 2016 (n = 78). Successful AVS required a selectivity index (cortisol adrenal vein /cortisol peripheral ) of ≥ 2.0 bilaterally. The overall success rate of C-arm CT-assisted AVS was 87%, which increased to 97.4% under QCA (P = .013). The procedure time (C-arm CT, 49.5 ± 21.3 min; QCA, 37.5 ± 15.6 min; P AVS. • Adrenal venous sampling (AVS) is a technically challenging procedure. • C-arm CT and quick cortisol assay (QCA) are efficient for assisting AVS. • QCA might outperform C-arm CT in enhancing AVS performance.

  10. Analysis of self-powered gamma ray detector with directional discrimination

    International Nuclear Information System (INIS)

    Levert, F.E.; Beyerlein, R.A.; Cox, S.A.

    1979-01-01

    The results of a combined Monte Carlo simulation and experimental investigation of the directional and energy dependent response of a self-powered gamma detector with a flat plate Pb-C central electrode are presented. The electron yield of the central electrode in a three dimensional mockup of the detector was calculated for photons of several discrete energies, emanating from an infinitely thin planar source, incident on the outer surface of the detector. Separate computations were done with the source facing the lead side and carbon side of the central electrode. Experimental measurements with a detector that closely matched the design used in the simulation were conducted in a graphite column next to a neutron leakage face of a low flux reactor. A localized gamma ray source was created by positioning a 235 U strip between the leakage face of the reactor and the detector. A comparison of results obtained in both cases showed good agreement. Also experimental measurements to determine the effect of the thickness of lead shielding surrounding the outer wall of the detector and space charge in the vacuum insulator between the central electrode and the inner wall on the response of the detector were performed. (Auth.)

  11. SU-D-204-05: Quantitative Comparison of a High Resolution Micro-Angiographic Fluoroscopic (MAF) Detector with a Standard Flat Panel Detector (FPD) Using the New Metric of Generalized Measured Relative Object Detectability (GM-ROD)

    Energy Technology Data Exchange (ETDEWEB)

    Russ, M; Ionita, C; Bednarek, D; Rudin, S [Toshiba Stroke and Vascular Research Center, University at Buffalo (SUNY), Buffalo, NY (United States)

    2015-06-15

    Purpose: In endovascular image-guided neuro-interventions, visualization of fine detail is paramount. For example, the ability of the interventionist to visualize the stent struts depends heavily on the x-ray imaging detector performance. Methods: A study to examine the relative performance of the high resolution MAF-CMOS (pixel size 75µm, Nyquist frequency 6.6 cycles/mm) and a standard Flat Panel Detector (pixel size 194µm, Nyquist frequency 2.5 cycles/mm) detectors in imaging a neuro stent was done using the Generalized Measured Relative Object Detectability (GM-ROD) metric. Low quantum noise images of a deployed stent were obtained by averaging 95 frames obtained by both detectors without changing other exposure or geometric parameters. The square of the Fourier transform of each image is taken and divided by the generalized normalized noise power spectrum to give an effective measured task-specific signal-to-noise ratio. This expression is then integrated from 0 to each of the detector’s Nyquist frequencies, and the GM-ROD value is determined by taking a ratio of the integrals for the MAF-CMOS to that of the FPD. The lower bound of integration can be varied to emphasize high frequencies in the detector comparisons. Results: The MAF-CMOS detector exhibits vastly superior performance over the FPD when integrating over all frequencies, yielding a GM-ROD value of 63.1. The lower bound of integration was stepped up in increments of 0.5 cycles/mm for higher frequency comparisons. As the lower bound increased, the GM-ROD value was augmented, reflecting the superior performance of the MAF-CMOS in the high frequency regime. Conclusion: GM-ROD is a versatile metric that can provide quantitative detector and task dependent comparisons that can be used as a basis for detector selection. Supported by NIH Grant: 2R01EB002873 and an equipment grant from Toshiba Medical Systems Corporation.

  12. In Silico Study of Spacer Arm Length Influence on Drug Vectorization by Fullerene C60

    Directory of Open Access Journals (Sweden)

    Haifa Khemir

    2015-01-01

    Full Text Available This work studies theoretically the effect of spacer arm lengths on the characteristics of a fullerene C60-based nanovector. The spacer arm is constituted of a carbon chain including a variable number of methylene groups (n = 2–11. To improve the ability of the fullerene carriage, two arms are presented simultaneously through a malonyl bridge. Then the evolution of selected physicochemical parameters is monitored as a function of the spacer arm length and the angle between the two arms. We show here that while the studied characteristics are almost independent of the spacer arm length or vary monotonically with it, the dipole moment and its orientation vary periodically with the parity of the number of carbon atoms. This periodicity is related to both modules and orientations of dipole moments of the spacer arms. In the field of chemical synthesis, these results highlight the importance of theoretical calculations for the optimization of operating conditions. In the field of drug discovery, they show that theoretical calculations of the chemical properties of a drug candidate can help predict its in vivo behaviour, notably its bioavailability and biodistribution, which are known to be tightly dependent of its polarity.

  13. Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction

    International Nuclear Information System (INIS)

    Wang, Adam S; Stayman, J Webster; Otake, Yoshito; Siewerdsen, Jeffrey H; Kleinszig, Gerhard; Vogt, Sebastian; Gallia, Gary L; Khanna, A Jay

    2014-01-01

    The potential for statistical image reconstruction methods such as penalized-likelihood (PL) to improve C-arm cone-beam CT (CBCT) soft-tissue visualization for intraoperative imaging over conventional filtered backprojection (FBP) is assessed in this work by making a fair comparison in relation to soft-tissue performance. A prototype mobile C-arm was used to scan anthropomorphic head and abdomen phantoms as well as a cadaveric torso at doses substantially lower than typical values in diagnostic CT, and the effects of dose reduction via tube current reduction and sparse sampling were also compared. Matched spatial resolution between PL and FBP was determined by the edge spread function of low-contrast (∼40–80 HU) spheres in the phantoms, which were representative of soft-tissue imaging tasks. PL using the non-quadratic Huber penalty was found to substantially reduce noise relative to FBP, especially at lower spatial resolution where PL provides a contrast-to-noise ratio increase up to 1.4–2.2× over FBP at 50% dose reduction across all objects. Comparison of sampling strategies indicates that soft-tissue imaging benefits from fully sampled acquisitions at dose above ∼1.7 mGy and benefits from 50% sparsity at dose below ∼1.0 mGy. Therefore, an appropriate sampling strategy along with the improved low-contrast visualization offered by statistical reconstruction demonstrates the potential for extending intraoperative C-arm CBCT to applications in soft-tissue interventions in neurosurgery as well as thoracic and abdominal surgeries by overcoming conventional tradeoffs in noise, spatial resolution, and dose. (paper)

  14. Novel Tetra-Primer ARMS-PCR Assays for Thiopurine Intolerance Susceptibility Mutations NUDT15 c.415C>T and TPMT c.719A>G (TPMT*3C in East Asians

    Directory of Open Access Journals (Sweden)

    Chi-Chun Ho

    2017-10-01

    Full Text Available Thiopurines are clinically useful in the management of diverse immunological and malignant conditions. Nevertheless, these purine analogues can cause lethal myelosuppression, which may be prevented by prospective testing for variants in the thiopurine S-methyltransferase (TPMT and, in East Asians, Nudix hydrolase 15 (NUDT15 genes. Two single-tube, tetra-primer amplification refractory mutation system polymerase chain reaction (ARMS-PCR assays were developed to genotype the common loss-of-function variants NUDT15 c.415C>T (rs116855232 and TPMT*3C c.719A>G (rs1142345. In a group of 60 unselected patients, one and seven were found to be homozygous and heterozygous, respectively, for NUDT15 c.415C>T; one was found to be heterozygous for TPMT*3C c.719A>G. There was no non-specific amplification, and the genotypes were 100% concordant with Sanger sequencing. Limit-of-detection for both assays was below 1 ng of heterozygous template per reaction. Time- and cost-effective ARMS-PCR assays, suitable for genotyping East-Asian patients for thiopurine intolerance, were successfully developed and validated.

  15. Analysis of patients' X-ray exposure in 146 percutaneous radiologic gastrostomies; Analyse der Strahlenexposition fuer Patienten bei 146 perkutanen radiologischen Gastrostomien

    Energy Technology Data Exchange (ETDEWEB)

    Petersen, Tim-Ole; Reinhardt, Martin; Fuchs, Jochen; Gosch, Dieter; Surov, Alexey; Stumpp, Patrick; Kahn, Thomas; Moche, Michael [Univ. Hospital Leipzig (Germany). Dept. of Diagnostic and Interventional Radiology

    2017-09-15

    intervention. Due to the significantly higher X-ray exposure, the use of a CBCT as well as PRG-placements using CT should be limited to clinically absolutely necessary exceptions with strict indication. Key Points: Fluoroscopically-guided PRG placements are interventions with low X-ray exposure. X-ray exposure from fluoroscopy is lower using flat panel detector systems as compared to image intensifier systems. The concomitant placement of an oro- or nasogastric probe extends the fluoroscopy time. Gastric probe placement is worthwhile to prevent the premature use of the significantly radiation-intensive CT. The use of the C-arm CT or the CT increases the beam exposure by 26 or 33 times, respectively. The PRG placement using C-arm CT and CT should only be performed in exceptional cases.

  16. Scatter correction using a primary modulator on a clinical angiography C-arm CT system.

    Science.gov (United States)

    Bier, Bastian; Berger, Martin; Maier, Andreas; Kachelrieß, Marc; Ritschl, Ludwig; Müller, Kerstin; Choi, Jang-Hwan; Fahrig, Rebecca

    2017-09-01

    Cone beam computed tomography (CBCT) suffers from a large amount of scatter, resulting in severe scatter artifacts in the reconstructions. Recently, a new scatter correction approach, called improved primary modulator scatter estimation (iPMSE), was introduced. That approach utilizes a primary modulator that is inserted between the X-ray source and the object. This modulation enables estimation of the scatter in the projection domain by optimizing an objective function with respect to the scatter estimate. Up to now the approach has not been implemented on a clinical angiography C-arm CT system. In our work, the iPMSE method is transferred to a clinical C-arm CBCT. Additional processing steps are added in order to compensate for the C-arm scanner motion and the automatic X-ray tube current modulation. These challenges were overcome by establishing a reference modulator database and a block-matching algorithm. Experiments with phantom and experimental in vivo data were performed to evaluate the method. We show that scatter correction using primary modulation is possible on a clinical C-arm CBCT. Scatter artifacts in the reconstructions are reduced with the newly extended method. Compared to a scan with a narrow collimation, our approach showed superior results with an improvement of the contrast and the contrast-to-noise ratio for the phantom experiments. In vivo data are evaluated by comparing the results with a scan with a narrow collimation and with a constant scatter correction approach. Scatter correction using primary modulation is possible on a clinical CBCT by compensating for the scanner motion and the tube current modulation. Scatter artifacts could be reduced in the reconstructions of phantom scans and in experimental in vivo data. © 2017 American Association of Physicists in Medicine.

  17. The value of flat-detector computed tomography during catheterisation of congenital heart disease

    Energy Technology Data Exchange (ETDEWEB)

    Gloeckler, Martin [University Hospital Erlangen, Department of Pediatric Cardiology, Erlangen (Germany); Friedrich-Alexander University Erlangen-Nuernberg, Department of Pediatric Cardiology, Erlangen (Germany); Koch, Andreas; Greim, Verena; Shabaiek, Amira; Dittrich, Sven [University Hospital Erlangen, Department of Pediatric Cardiology, Erlangen (Germany); Rueffer, Andre; Cesnjevar, Robert [University Hospital Erlangen, Department of Congenital Heart Surgery, Erlangen (Germany); Achenbach, Stephan [University Hospital Erlangen, Department of Cardiology, Erlangen (Germany)

    2011-12-15

    To analyse the diagnostic utility of flat-detector computed tomography imaging (FD-CT) in patients with congenital heart disease, including the value of image fusion to overlay three-dimensional (3D) reconstructions on fluoroscopic images during catheter-based interventions. We retrospectively analysed 62 consecutive paediatric patients in whom FD-CT was used during catheterisation of congenital heart disease. Expert operators rated the clinical value of FD-CT over conventional fluoroscopic imaging. Added radiation exposure and contrast medium volume were evaluated. During a 12-month period, FD-CT was performed in 62 out of 303 cardiac catheterisations. Median patient age was 3.5 years. In 32/62 cases, FD-CT was used for diagnostic purposes, in 30/62 cases it was used in the context of interventions. Diagnostic utility was never rated as ''misleading''. It was classified as ''not useful'' in six cases (9.7%), ''useful'' in 18 cases (29.0%), ''very useful'' in 37 cases (59.7%) and ''essential'' in one case (1.6%). The median added dose-area product was 111.0 {mu}Gym{sup 2}, the required additional quantity of contrast medium was 1.6 ml/kg. FD-CT provides useful diagnostic information in most of the patients investigated for congenital heart disease. The added radiation exposure and contrast medium volume are reasonable. (orig.)

  18. Technical characterization of five x-ray detectors for paediatric radiography applications

    Science.gov (United States)

    Marshall, N. W.; Smet, M.; Hofmans, M.; Pauwels, H.; De Clercq, T.; Bosmans, H.

    2017-12-01

    Physical image quality of five x-ray detectors used in the paediatric imaging department is characterized with the aim of establishing the range/scope of imaging performance provided by these detectors for neonatal imaging. Two computed radiography (CR) detectors (MD4.0 powder imaging plate (PIP) and HD5.0 needle imaging plate (NIP), Agfa HealthCare NV, B-2640 Mortsel, Belgium) and three flat panel detectors (FPD) (the Agfa DX-D35C and DX-D45C and the DRX-2530C (Carestream Health Inc., Rochester, NY 14608, USA)) were assessed. Physical image quality was characterized using the detector metrics given by the International Electrotechnical Commission (IEC 62220-1) to measure modulation transfer function (MTF), the noise power spectrum (NPS) and the detective quantum efficiency (DQE) using the IEC-specified beam qualities of RQA3 and RQA5. The DQE was evaluated at the normal operating detector air kerma (DAK) level, defined at 2.5 µGy for all detectors, and at factors of 1/3.2 and 3.2 times the normal level. MTF curves for the different detectors were similar at both RQA3 and RQA5 energies; the average spatial frequency for the 50% point (MTF0.5) at RQA3 was 1.26 mm-1, with a range from 1.20 mm-1 to 1.37 mm-1. The DQE of the NIP CR compared to the PIP CR was notably greater and similar to that for the FPD devices. At RQA3, average DQE for the FPD and NIP (at 0.5 mm-1 2.5 µGy) was 0.57 compared to 0.26 for the PIP CR. At the RQA5 energy, the DRX-2530C and the DX-D45C had the highest DQE (~0.6 at 0.5 mm-1 2.5 µGy). Noise separation analysis using the polynomial model showed higher electronic noise for the DX-D35C and DRX-2530C detectors; this explains the reduced DQE seen at 0.7 µGy/image. The NIP CR detector offers notably improved DQE performance compared to the PIP CR system and a value similar to the DQE for FPD devices at the RQA3 energy.

  19. Optimized intravenous flat detector CT for non-invasive visualization of intracranial stents: first results

    International Nuclear Information System (INIS)

    Struffert, Tobias; Kloska, Stephan; Engelhorn, Tobias; Ott, Sabine; Doelken, Marc; Saake, Marc; Doerfler, Arnd; Deuerling-Zheng, Yu.; Koehrmann, Martin

    2011-01-01

    As stents for treating intracranial atherosclerotic stenosis may develop in-stent re-stenosis (ISR) in up to 30%, follow-up imaging is mandatory. Residual stenosis (RS) is not rare. We evaluated an optimised Flat Detector CT protocol with intravenous contrast material application (i.v. FD-CTA) for non-invasive follow-up. In 12 patients with intracranial stents, follow-up imaging was performed using i.v. FD-CTA. MPR, subtracted MIP and VRT reconstructions were used to correlate to intra-arterial angiography (DSA). Two neuroradiologists evaluated the images in anonymous consensus reading and calculated the ISR or RS. Correlation coefficients and a Wilcoxon test were used for statistical analysis. In 4 patients, no stenosis was detected. In 6 patients RS and in two cases ISR by intima hyperplasia perfectly visible on MPR reconstructions of i.v. FD-CTA were detected. Wilcoxon's test showed no significant differences between the methods (p > 0.05). We found a high correlation with coefficients of the pairs DSA/ FD-CT MIP r = 0.91, DSA/ FD-CT MPR r = 0.82 and FD-CT MIP/ FD-CT MPR r = 0.8. Intravenous FD-CTA could clearly visualise the stent and the lumen, allowing ISR or RS to be recognised. FD-CTA provides a non-invasive depiction of intracranial stents and might replace DSA for non-invasive follow-up imaging. (orig.)

  20. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality

    Science.gov (United States)

    Vano, E.; Geiger, B.; Schreiner, A.; Back, C.; Beissel, J.

    2005-12-01

    The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 µGy/frame (cine) and 5 and 95 mGy min-1 (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality.

  1. Blackbody radiation from light cone in flat space time

    International Nuclear Information System (INIS)

    Gerlach, U.H.

    1983-01-01

    Blackbody radiation in flat space-time is not necessarily associated with the flat event horizon of a single accelerated observer. The author considers a spherical bubble which expands in a uniformly accelerating fashion. Its history traces out a time-like hyperboloid. Suppose the bubble membrane has a spatially isotropic and homogeneous (surface) stress energy tensor i.e. the membrane is made out of the stiffest possible material permitted by causality considerations. It follows that this bubble membrane is in equilibrium even though it is expanding. Such an expanding bubble membrane may serve as a detector of electromagnetic radiation if the membrane can interact with the electromagnetic field. (Auth.)

  2. Characteristics of Fabricated SiC Neutron Detectors for Neutron Flux Monitoring

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Han Soo; Ha, Jang Ho; Park, Se Hwan; Lee, Kyu Hong [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of); Lee, Cheol Ho [Hanyang University, Seoul (Korea, Republic of)

    2011-05-15

    An SPND (Self-powered Neutron Detector) is commonly used for neutron detection in NPP (Nuclear Power Plant) by virtue of un-reactivity for gamma-rays. But it has a drawback, which is that it cannot detect neutrons in real time due to beta emissions (about > 48 s) after reactions between neutrons and {sup 103}Rh in an SPND. And Generation IV reactors such as MSR (Molten-salt reactor), SFR (Sodium-cooled fast reactor), and GFR (Gas-cooled fast reactor) are designed to compact size and integration type. For GEN IV reactor, neutron monitor also must be compact-sized to apply such reactor easily and much more reliable. The wide band-gap semiconductors such as SiC, AlN, and diamond make them an attractive alternative in applications in harsh environments by virtue of the lower operating voltage, faster charge-collection times compared with gas-filled detectors, and compact size.1) In this study, two PIN-type SiC semiconductor neutron detectors, which are for fast neutron detection by elastic and inelastic scattering SiC atoms and for thermal neutron detection by charged particle emissions of 6LiF reaction, were designed and fabricated for NPP-related applications. Preliminary tests such as I-V and alpha response were performed and neutron responses at ENF in HANARO research reactor were also addressed. The application feasibility of the fabricated SiC neutron detector as an in-core neutron monitor was discussed

  3. Novel Heterongineered Detectors for Multi-Color Infrared Sensing

    Science.gov (United States)

    2012-01-30

    a) Sequential but collocated two-color detection capabilities of type II InAsGaSb SLS detector based on an nBn design and (b) Simultaneous and...captions: Figure 1. Heterostructure Schematic of (a) PbIbN Design, (b) Dual Color Detector Design with flat band energy lineups . Figure 2. (a) Spectral

  4. Development of CT and 3D-CT Using Flat Panel Detector Based Real-Time Digital Radiography System

    International Nuclear Information System (INIS)

    Ravindran, V. R.; Sreelakshmi, C.; Vibin

    2008-01-01

    The application of Digital Radiography in the Nondestructive Evaluation (NDE) of space vehicle components is a recent development in India. A Real-time DR system based on amorphous silicon Flat Panel Detector has been developed for the NDE of solid rocket motors at Rocket Propellant Plant of VSSC in a few years back. The technique has been successfully established for the nondestructive evaluation of solid rocket motors. The DR images recorded for a few solid rocket specimens are presented in the paper. The Real-time DR system is capable of generating sufficient digital X-ray image data with object rotation for the CT image reconstruction. In this paper the indigenous development of CT imaging based on the Realtime DR system for solid rocket motor is presented. Studies are also carried out to generate 3D-CT image from a set of adjacent CT images of the rocket motor. The capability of revealing the spatial location and characterisation of defect is demonstrated by the CT and 3D-CT images generated.

  5. Percutaneous radiofrequency ablation of lung metastases from colorectal carcinoma under C-arm cone beam CT guidance.

    Science.gov (United States)

    Amouyal, G; Pernot, S; Déan, C; Cholley, B; Scotté, F; Sapoval, M; Pellerin, O

    2017-11-01

    The aim of this study was to assess the feasibility, safety and efficacy of percutaneous radiofrequency ablation of lung metastases from colorectal carcinoma using C-arm cone beam computed tomography (CBCT) guidance. This single-center prospective observational study was performed from August 2013 to August 2016, and included consecutive patients referred for radiofrequency ablation of lung metastases from colorectal cancer. Radiofrequency ablation procedures were performed under C-arm CBCT guidance. Feasibility was assessed by probe accuracy placement, time to accurate placement and number of C-arm CBCT acquisitions to reach the target lesion. Safety was assessed by the report of adverse event graded using the common terminology criteria for adverse events (CTCAE-V4.0). Efficacy was assessed by metastases response rate using RECIST 1.1 and 18 FDG-PET-CT tumor uptake at 6months. Fifty-four consecutive patients (32 men, 22 women) with a mean age of 63±8 (SD) years (range: 51-81years) with a total of 56 lung metastasis from colorectal metastases were treated in a single session. The mean tumor diameter was 25.6±4.5 (SD)mm (range: 17-31mm). Median time to insert the needle into the target lesion was 10min (range: 5-25min). Median number of needles repositioning and C-arm CBCT acquisition per patient was 1 (range: 0-3) and 4 (range: 3-6) respectively. The accuracy for radiofrequency ablation probe placement was 2±0.2 (SD)mm (range: 0-9mm). Pneumothorax requiring chest tube placement occurred in one patient (CTCAE-V4.0 grade 3). At 6months, all patients were alive with tumor response rate of -27% and had no significant activity on the 18 FDG-PET CT follow-up. Percutaneous radiofrequency ablation of lung metastases from colorectal cancer under C-arm CBCT guidance is feasible and safe, with immediate and short-term results similar to those obtained using conventional CT guidance. Copyright © 2017 Éditions françaises de radiologie. Published by Elsevier Masson SAS

  6. Three-dimensional C-arm CT-guided transjugular intrahepatic portosystemic shunt placement: Feasibility, technical success and procedural time

    Energy Technology Data Exchange (ETDEWEB)

    Ketelsen, Dominik; Groezinger, Gerd; Maurer, Michael; Grosse, Ulrich; Horger, Marius; Nikolaou, Konstantin; Syha, Roland [University of Tuebingen, Department of Diagnostic and Interventional Radiology, Tuebingen (Germany); Lauer, Ulrich M. [University of Tuebingen, Internal Medicine I, Department of Gastroenterology, Hepatology and Infectious disease, Tuebingen (Germany)

    2016-12-15

    Establishment of transjugular intrahepatic portosystemic shunts (TIPS) constitutes a standard procedure in patients suffering from portal hypertension. The most difficult step in TIPS placement is blind puncture of the portal vein. This study aimed to evaluate three-dimensional mapping of portal vein branches and targeted puncture of the portal vein. Twelve consecutive patients suffering from refractory ascites by liver cirrhosis were included in this retrospective study to evaluate feasibility, technical success and procedural time of C-arm CT-targeted puncture of the portal vein. As a control, 22 patients receiving TIPS placement with fluoroscopy-guided blind puncture were included to compare procedural time. Technical success could be obtained in 100 % of the study group (targeted puncture) and in 95.5 % of the control group (blind puncture). Appropriate, three-dimensional C-arm CT-guided mapping of the portal vein branches could be achieved in all patients. The median number of punctures in the C-arm CT-guided study group was 2 ± 1.3 punctures. Procedural time was significantly lower in the study group (14.8 ± 8.2 min) compared to the control group (32.6 ± 22.7 min) (p = 0.02). C-arm CT-guided portal vein mapping is technically feasible and a promising tool for TIPS placement resulting in a significant reduction of procedural time. (orig.)

  7. Engineering flat electronic bands in quasiperiodic and fractal loop geometries

    Energy Technology Data Exchange (ETDEWEB)

    Nandy, Atanu, E-mail: atanunandy1989@gmail.com; Chakrabarti, Arunava, E-mail: arunava_chakrabarti@yahoo.co.in

    2015-11-06

    Exact construction of one electron eigenstates with flat, non-dispersive bands, and localized over clusters of various sizes is reported for a class of quasi-one-dimensional looped networks. Quasiperiodic Fibonacci and Berker fractal geometries are embedded in the arms of the loop threaded by a uniform magnetic flux. We work out an analytical scheme to unravel the localized single particle states pinned at various atomic sites or over clusters of them. The magnetic field is varied to control, in a subtle way, the extent of localization and the location of the flat band states in energy space. In addition to this we show that an appropriate tuning of the field can lead to a re-entrant behavior of the effective mass of the electron in a band, with a periodic flip in its sign. - Highlights: • Exact construction of eigenstates with flat and dispersive bands is reported. • Competition between translational order and growth of aperiodicity is discussed. • The effect of magnetic field on the location of flat band states is shown. • Flux tunable re-entrant behavior of the effective mass of electron is studied.

  8. A new parameter in the electrochemical etching of polymer track detectors

    International Nuclear Information System (INIS)

    Sohrabi, M.; Katouzi, M.

    1993-01-01

    It was discovered that the pressure applied to the electrochemical etching (ECE) chamber system and in turn to washers holding the detector tight in place between two semi-chambers has a direct effect on the internal heating and time to breakdown of the polymer detector. The effect was found to be dependent on the type, material, shape and size of the washers holding the detector in place under pressure. To verify such parameters, a pressure ECE chamber (PECE) with measurable and reproducible pressure was designed and constructed. Three types of rubber washers, such as ''O'' rings, flat rings and sheets as well as polycarbonate (PC) detectors glued directly between two semi-syringes, were used. Flat rubber sheets were shown to have relatively minor effects on the internal heating rate and are recommended. The effect seems to be due to forced vibrations of the detector under an electric field, the frequency of which depends on the degree to which the detector is stretched under pressure, like winding the strings of a musical instrument. The results of the above studies are presented and discussed. (orig.)

  9. A report on the status of the GEO 600 gravitational wave detector

    International Nuclear Information System (INIS)

    Hewitson, M; Aufmuth, P; Aulbert, C

    2003-01-01

    GEO 600 is an interferometric gravitational wave detector with 600 m arms, which will employ a novel, dual-recycled optical scheme allowing its optical response to be tuned over a range of frequencies (from ∼100 Hz to a few kHz). Additional advanced technologies, such as multiple pendulum suspensions with monolithic bottom stages, make the anticipated sensitivity of GEO 600 comparable to initial detectors with kilometre arm lengths. This paper discusses briefly the design of GEO, reports on the status of the detector up to the end of 2002 with particular focus on participation in coincident engineering and science runs with LIGO detectors. The plans leading to a fully optimized detector and participation in future coincident science runs are briefly outlined

  10. Improved Correction of IR Loss in Diffuse Shortwave Measurements: An ARM Value-Added Product

    Energy Technology Data Exchange (ETDEWEB)

    Younkin, K; Long, CN

    2003-11-01

    Simple single black detector pyranometers, such as the Eppley Precision Spectral Pyranometer (PSP) used by the Atmospheric Radiation Measurement (ARM) Program, are known to lose energy via infrared (IR) emission to the sky. This is especially a problem when making clear-sky diffuse shortwave (SW) measurements, which are inherently of low magnitude and suffer the greatest IR loss. Dutton et al. (2001) proposed a technique using information from collocated pyrgeometers to help compensate for this IR loss. The technique uses an empirically derived relationship between the pyrgeometer detector data (and alternatively the detector data plus the difference between the pyrgeometer case and dome temperatures) and the nighttime pyranometer IR loss data. This relationship is then used to apply a correction to the diffuse SW data during daylight hours. We developed an ARM value-added product (VAP) called the SW DIFF CORR 1DUTT VAP to apply the Dutton et al. correction technique to ARM PSP diffuse SW measurements.

  11. The thermal neutron detection using 4H-SiC detectors with 6LiF conversion layer

    International Nuclear Information System (INIS)

    Zatko, B.; Bohacek, P.; Sekacova, M.; Arbet, J.; Sagatova, A.; Necas, V.

    2016-01-01

    In this paper we have examined 4H-SiC detector using a thermal neutron source and studied its detection properties. The detector was exposed to neutrons generated by 238 Pu-Be radiation source. The detection properties of 4H-SiC detectors were evaluated considering the use of the 6 LiF conversion. We prepared 4H-SiC Schottky contact detectors based on high-quality of epitaxial layer. The current-voltage characteristic show operating region between 100 V and 400 V. The detector was connected to the spectrometric set-up and used for detection of alpha particles from 241 Am. Following the 6 LiF conversion layer was applied on the Schottky contact of detector and the detection of thermal neutrons was performed. We are able to resolve alpha particles and tritons which are products of nuclear reaction between thermal neutrons and conversion layer. Also bare detector was used for neutron detection to clearly show significant influence of the used conversion layer.(authors)

  12. Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. II. Experiment

    International Nuclear Information System (INIS)

    Choi, Jang-Hwan; Maier, Andreas; Keil, Andreas; McWalter, Emily J.; Gold, Garry E.; Fahrig, Rebecca; Pal, Saikat; Beaupré, Gary S.

    2014-01-01

    Purpose: A C-arm CT system has been shown to be capable of scanning a single cadaver leg under loaded conditions by virtue of its highly flexible acquisition trajectories. In Part I of this study, using the 4D XCAT-based numerical simulation, the authors predicted that the involuntary motion in the lower body of subjects in weight-bearing positions would seriously degrade image quality and the authors suggested three motion compensation methods by which the reconstructions could be corrected to provide diagnostic image quality. Here, the authors demonstrate that a flat-panel angiography system is appropriate for scanning both legs of subjectsin vivo under weight-bearing conditions and further evaluate the three motion-correction algorithms using in vivo data. Methods: The geometry of a C-arm CT system for a horizontal scan trajectory was calibrated using the PDS-2 phantom. The authors acquired images of two healthy volunteers while lying supine on a table, standing, and squatting at several knee flexion angles. In order to identify the involuntary motion of the lower body, nine 1-mm-diameter tantalum fiducial markers were attached around the knee. The static mean marker position in 3D, a reference for motion compensation, was estimated by back-projecting detected markers in multiple projections using calibrated projection matrices and identifying the intersection points in 3D of the back-projected rays. Motion was corrected using three different methods (described in detail previously): (1) 2D projection shifting, (2) 2D deformable projection warping, and (3) 3D rigid body warping. For quantitative image quality analysis, SSIM indices for the three methods were compared using the supine data as a ground truth. Results: A 2D Euclidean distance-based metric of subjects’ motion ranged from 0.85 mm (±0.49 mm) to 3.82 mm (±2.91 mm) (corresponding to 2.76 to 12.41 pixels) resulting in severe motion artifacts in 3D reconstructions. Shifting in 2D, 2D warping, and 3D

  13. Fast neutron detection at near-core location of a research reactor with a SiC detector

    Science.gov (United States)

    Wang, Lei; Jarrell, Josh; Xue, Sha; Tan, Chuting; Blue, Thomas; Cao, Lei R.

    2018-04-01

    The measurable charged-particle produced from the fast neutron interactions with the Si and C nucleuses can make a wide bandgap silicon carbide (SiC) sensor intrinsically sensitive to neutrons. The 4H-SiC Schottky detectors have been fabricated and tested at up to 500 °C, presenting only a slightly degraded energy resolution. The response spectrum of the SiC detectors were also obtained by exposing the detectors to external neutron beam irradiation and at a near-core location where gamma-ray field is intense. The fast neutron flux of these two locations are ∼ 4 . 8 × 104cm-2 ṡs-1 and ∼ 2 . 2 × 107cm-2 ṡs-1, respectively. At the external beam location, a Si detector was irradiated side-by-side with SiC detector to disjoin the neutron response from Si atoms. The contribution of gamma ray, neutron scattering, and charged-particles producing reactions in the SiC was discussed. The fast neutron detection efficiencies were determined to be 6 . 43 × 10-4 for the external fast neutron beam irradiation and 6 . 13 × 10-6 for the near-core fast neutron irradiation.

  14. Automation of the Characterization of High Purity Germanium Detectors

    Science.gov (United States)

    Dugger, Charles ``Chip''

    2014-09-01

    Neutrinoless double beta decay is a rare hypothesized process that may yield valuable insight into the fundamental properties of the neutrino. Currently there are several experiments trying to observe this process, including the Majorana DEMONSTRAOR experiment, which uses high purity germanium (HPGe) detectors to generate and search for these events. Because the event happens internally, it is essential to have the lowest background possible. This is done through passive detector shielding, as well as event discrimination techniques that distinguish between multi-site events characteristic of gamma-radiation, and single-site events characteristic of neutrinoless double beta decay. Before fielding such an experiment, the radiation response of the detectors must be characterized. A robotic arm is being tested for future calibration of HPGe detectors. The arm will hold a source at locations relative to the crystal while data is acquired. Several radioactive sources of varying energy levels will be used to determine the characteristics of the crystal. In this poster, I will present our work with the robot, as well as the characterization of data we took with an underground HPGe detector at the WIPP facility in Carlsbad, NM (2013). Neutrinoless double beta decay is a rare hypothesized process that may yield valuable insight into the fundamental properties of the neutrino. Currently there are several experiments trying to observe this process, including the Majorana DEMONSTRAOR experiment, which uses high purity germanium (HPGe) detectors to generate and search for these events. Because the event happens internally, it is essential to have the lowest background possible. This is done through passive detector shielding, as well as event discrimination techniques that distinguish between multi-site events characteristic of gamma-radiation, and single-site events characteristic of neutrinoless double beta decay. Before fielding such an experiment, the radiation response of

  15. PHENIX central arm tracking detectors

    International Nuclear Information System (INIS)

    Adcox, K.; Ajitanand, N.N.; Alexander, J.; Autrey, D.; Averbeck, R.; Azmoun, B.; Barish, K.N.; Baublis, V.V.; Belkin, R.; Bhaganatula, S.; Biggs, J.C.; Borland, D.; Botelho, S.; Bryan, W.L.; Burward-Hoy, J.; Butsyk, S.A.; Chang, W.C.; Christ, T.; Dietzsch, O.; Drees, A.; Rietz, R. du; El Chenawi, K.; Evseev, V.A.; Fellenstein, J.; Ferdousi, T.; Fraenkel, Z.; Franz, A.; Fung, S.Y.; Gannon, J.; Garpman, S.; Godoi, A.L.; Greene, S.V.; Gustafsson, H.-A.; Harder, J.; Hemmick, T.K.; Heuser, J.M.; Holzmann, W.; Hutter, R.; Issah, M.; Ivanov, V.I.; Jacak, B.V.; Jagadish, U.; Jia, J.; Johnson, S.C.; Kandasamy, A.; Kann, M.R.; Kelley, M.A.; Khanzadeev, A.V.; Khomutnikov, A.; Komkov, B.G.; Kopytine, M.L.; Kotchenda, L.; Kotchetkov, D.; Kozlov, V.S.; Kravtsov, P.A.; Kudin, L.G.; Kuriatkov, V.V.; Lacey, R.; Lauret, J.; Lebedev, A.; Lebedev, V.D.; Li, X.H.; Libby, B.; Liccardi, W.; Machnowski, R.; Mahon, J.; Markushin, D.G.; Matathias, F.; Marx, M.D.; Messer, F.; Miftakhov, N.M.; Milan, J.; Miller, T.E.; Milov, A.; Minuzzo, K.; Mioduszewski, S.; Mitchell, J.T.; Muniruzzamann, M.; Nandi, B.K.; Negrin, J.; Nilsson, P.; Nystrand, J.; O'Brien, E.; O'Connor, P.; Oskarsson, A.; Oesterman, L.; Otterlund, I.; Pancake, C.E.; Pantuev, V.S.; Petersen, R.; Pinkenburg, C.H.; Pisani, R.P.; Purwar, A.K.; Rankowitz, S.; Ravinovich, I.; Riabov, V.G.; Riabov, Yu.G.; Rosati, M.; Rose, A.A.; Roschin, E.V.; Samsonov, V.M.; Sangster, T.C.; Seto, R.; Silvermyr, D.; Sivertz, M.; Smith, M.; Solodov, G.P.; Stenlund, E.; Takagui, E.M.; Tarakanov, V.I.; Tarasenkova, O.P.; Thomas, J.L.; Trofimov, V.A.; Tserruya, I.; Tydesjoe, H.; Velkovska, J.; Velkovsky, M.; Vishnevskii, V.I.; Vorobyov, A.A.; Vznuzdaev, E.A.; Vznuzdaev, M.; Wang, H.Q.; Weimer, T.; Wolniewicz, K.; Wu, J.; Xie, W.; Young, G.R.

    2003-01-01

    The PHENIX tracking system consists of Drift Chambers (DC), Pad Chambers (PC) and the Time Expansion Chamber (TEC). PC1/DC and PC2/TEC/PC3 form the inner and outer tracking units, respectively. These units link the track segments that transverse the RICH and extend to the EMCal. The DC measures charged particle trajectories in the r-phi direction to determine p T of the particles and the invariant mass of particle pairs. The PCs perform 3D spatial point measurements for pattern recognition and longitudinal momentum reconstruction and provide spatial resolution of a few mm in both r-phi and z. The TEC tracks particles passing through the region between the RICH and the EMCal. The design and operational parameters of the detectors are presented and running experience during the first year of data taking with PHENIX is discussed. The observed spatial and momentum resolution is given which imposes a limitation on the identification and characterization of charged particles in various momentum ranges

  16. Dynamic defectoscopy with flat panel and CdTe Timepix X-ray detectors combined with an optical camera

    Czech Academy of Sciences Publication Activity Database

    Vavřík, Daniel; Fauler, A.; Fiederle, M.; Jandejsek, Ivan; Jakůbek, J.; Tureček, D.; Zwerger, A.

    2013-01-01

    Roč. 8, April (2013), C04009 ISSN 1748-0221. [International Workshop on Radiation Imaging Detectors /14./. Figueira da Foz, Coimbra, 01.07.2012-05.07.2012] R&D Projects: GA ČR(CZ) GA103/09/2101 Institutional support: RVO:68378297 Keywords : X-ray digital radiography * fracture mechanics * crack path * X-ray defectoscopy Subject RIV: JM - Building Engineering Impact factor: 1.526, year: 2013 http://iopscience.iop.org/1748-0221/8/04/C04009/

  17. Hydraulic actuators for flexible robots : a flatness based approach for tracking and vibration control

    NARCIS (Netherlands)

    Wey, T.; Lemmen, M.; Bernzen, W.; Wey, T.

    1999-01-01

    This paper deals with an application of the differential algebraic flatness approach to hydraulic drives. Here, an elastic robot arm driven by a differential cylinder is investigated. The task is to design a suitable control law which not only tracks a given trajectory but also allows the damping of

  18. Flat deformation theorem and symmetries in spacetime

    International Nuclear Information System (INIS)

    Llosa, Josep; Carot, Jaume

    2009-01-01

    The flat deformation theorem states that given a semi-Riemannian analytic metric g on a manifold, locally there always exists a two-form F, a scalar function c, and an arbitrarily prescribed scalar constraint depending on the point x of the manifold and on F and c, say Ψ(c, F, x) = 0, such that the deformed metric η = cg - εF 2 is semi-Riemannian and flat. In this paper we first show that the above result implies that every (Lorentzian analytic) metric g may be written in the extended Kerr-Schild form, namely η ab := ag ab - 2bk (a l b) where η is flat and k a , l a are two null covectors such that k a l a = -1; next we show how the symmetries of g are connected to those of η, more precisely; we show that if the original metric g admits a conformal Killing vector (including Killing vectors and homotheties), then the deformation may be carried out in a way such that the flat deformed metric η 'inherits' that symmetry.

  19. Rotating detectors and Mach's principle

    Energy Technology Data Exchange (ETDEWEB)

    Paola, R.D.M. de; Svaiter, N.F

    2000-08-01

    In this work we consider a quantum version of Newton{sup s} bucket experiment in a fl;at spacetime: we take an Unruh-DeWitt detector in interaction with a real massless scalar field. We calculate the detector's excitation rate when it is uniformly rotating around some fixed point and the field is prepared in the Minkowski vacuum and also when the detector is inertial and the field is in the Trocheries-Takeno vacuum state. These results are compared and the relations with Mach's principle are discussed. (author)

  20. Image quality and localization accuracy in C-arm tomosynthesis-guided head and neck surgery

    International Nuclear Information System (INIS)

    Bachar, G.; Siewerdsen, J. H.; Daly, M. J.; Jaffray, D. A.; Irish, J. C.

    2007-01-01

    The image quality and localization accuracy for C-arm tomosynthesis and cone-beam computed tomography (CBCT) guidance of head and neck surgery were investigated. A continuum in image acquisition was explored, ranging from a single exposure (radiograph) to multiple projections acquired over a limited arc (tomosynthesis) to a full semicircular trajectory (CBCT). Experiments were performed using a prototype mobile C-arm modified to perform 3D image acquisition (a modified Siemens PowerMobil). The tradeoffs in image quality associated with the extent of the source-detector arc (θ tot ), the number of projection views, and the total imaging dose were evaluated in phantom and cadaver studies. Surgical localization performance was evaluated using three cadaver heads imaged as a function of θ tot . Six localization tasks were considered, ranging from high-contrast feature identification (e.g., tip of a K-wire pointer) to more challenging soft-tissue delineation (e.g., junction of the hard and soft palate). Five head and neck surgeons and one radiologist participated as observers. For each localization task, the 3D coordinates of landmarks pinpointed by each observer were analyzed as a function of θ tot . For all tomosynthesis angles, image quality was highest in the coronal plane, whereas sagittal and axial planes exhibited a substantial decrease in spatial resolution associated with out-of-plane blur and distortion. Tasks involving complex, lower-contrast features demonstrated steeper degradation with smaller tomosynthetic arc. Localization accuracy in the coronal plane was correspondingly high, maintained to tot ∼30 deg. , whereas sagittal and axial localization degraded rapidly below θ tot ∼60 deg. . Similarly, localization precision was better than ∼1 mm within the coronal plane, compared to ∼2-3 mm out-of-plane for tomosynthesis angles below θ tot ∼45 deg. . An overall 3D localization accuracy of ∼2.5 mm was achieved with θ tot ∼ 90 deg. for most

  1. X-ray light valve (XLV): a novel detectors' technology for digital mammography

    Science.gov (United States)

    Marcovici, Sorin; Sukhovatkin, Vlad; Oakham, Peter

    2014-03-01

    A novel method, based on X-ray Light Valve (XLV) technology, is proposed for making good image quality yet inexpensive flat panel detectors for digital mammography. The digital mammography markets, particularly in the developing countries, demand quality machines at substantially lower prices than the ones available today. Continuous pressure is applied on x-ray detectors' manufacturers to reduce the flat panel detectors' prices. XLV presents a unique opportunity to achieve the needed price - performance characteristics for direct conversion, x-ray detectors. The XLV based detectors combine the proven, superior, spatial resolution of a-Se with the simplicity and low cost of liquid crystals and optical scanning. The x-ray quanta absorbed by a 200 μm a-Se produce electron - hole pairs that move under an electric field to the top and bottom of a-Se layer. This 2D charge distribution creates at the interface with the liquid crystals a continuous (analog) charge image corresponding to the impinging radiation's information. Under the influence of local electrical charges next to them, the liquid crystals twist proportionally to the charges and vary their light reflectivity. A scanning light source illuminates the liquid crystals while an associated, pixilated photo-detector, having a 42 μm pixel size, captures the light reflected by the liquid crystals and converts it in16 bit words that are transmitted to the machine for image processing and display. The paper will describe a novel XLV, 25 cm x 30 cm, flat panel detector structure and its underlying physics as well as its preliminary performance measured on several engineering prototypes. In particular, the paper will present the results of measuring XLV detectors' DQE, MTF, dynamic range, low contrast resolution and dynamic behavior. Finally, the paper will introduce the new, low cost, XLV detector based, digital mammography machine under development at XLV Diagnostics Inc.

  2. SiC detectors to monitor ionizing radiations emitted from nuclear events and plasmas

    Science.gov (United States)

    Torrisi, L.; Cannavò, A.

    2016-09-01

    Silicon Carbide (SiC) semiconductor detectors are increasingly employed in Nuclear Physics for their advantages with respect to traditional silicon (Si). Such detectors show an energy resolution, charge mobility, response velocity and detection efficiency similar to Si detectors. However, the higher band gap (3.26 eV), the lower leakage current (∼10 pA) maintained also at room temperature, the higher radiation hardness and the higher density with respect to Si represent some indisputable advantages characterizing such detectors. The devices can be employed at high temperatures, at high absorbed doses and in the case of high visible light intensities, for example, in plasma, for limited exposition times without damage. Generally SiC Schottky diodes are employed in reverse polarization with an active region depth of the order of 100 µm, purity below 1014 cm-3 and an active area lower than 1 cm2. Measurements in the regime of proportionality with the radiation energy released in the active region and measurements in time-of-flight configuration are employed for nuclear emission events produced at both low and high fluences. Alpha spectra demonstrated an energy resolution of about 1.3% at 5.8 MeV. Radiation emission from laser-generated plasma can be monitored in terms of detected photons, electrons and ions, using the laser pulse as a start signal and the radiation detection as a stop signal, enabling to measure the ion velocity by knowing the target-detector flight distance. SiC spectra acquired in the Messina University laboratories using radioactive ion sources and at the PALS laboratory facility in Prague (Czech Republic) are presented. A preliminary study of the use of SiC detectors, embedded in a water equivalent polymer, as a dosimeter is presented and discussed.

  3. Use and imaging performance of CMOS flat panel imager with LiF/ZnS(Ag) and Gadox scintillation screens for neutron radiography

    Science.gov (United States)

    Cha, B. K.; kim, J. Y.; Kim, T. J.; Sim, C.; Cho, G.; Lee, D. H.; Seo, C.-W.; Jeon, S.; Huh, Y.

    2011-01-01

    In digital neutron radiography system, a thermal neutron imaging detector based on neutron-sensitive scintillating screens with CMOS(complementary metal oxide semiconductor) flat panel imager is introduced for non-destructive testing (NDT) application. Recently, large area CMOS APS (active-pixel sensor) in conjunction with scintillation films has been widely used in many digital X-ray imaging applications. Instead of typical imaging detectors such as image plates, cooled-CCD cameras and amorphous silicon flat panel detectors in combination with scintillation screens, we tried to apply a scintillator-based CMOS APS to neutron imaging detection systems for high resolution neutron radiography. In this work, two major Gd2O2S:Tb and 6LiF/ZnS:Ag scintillation screens with various thickness were fabricated by a screen printing method. These neutron converter screens consist of a dispersion of Gd2O2S:Tb and 6LiF/ZnS:Ag scintillating particles in acrylic binder. These scintillating screens coupled-CMOS flat panel imager with 25x50mm2 active area and 48μm pixel pitch was used for neutron radiography. Thermal neutron flux with 6x106n/cm2/s was utilized at the NRF facility of HANARO in KAERI. The neutron imaging characterization of the used detector was investigated in terms of relative light output, linearity and spatial resolution in detail. The experimental results of scintillating screen-based CMOS flat panel detectors demonstrate possibility of high sensitive and high spatial resolution imaging in neutron radiography system.

  4. Particle gamma correlations in {sup 12}C measured with the CsI(Tl) based detector array CHIMERA

    Energy Technology Data Exchange (ETDEWEB)

    Cardella, G., E-mail: cardella@ct.infn.it [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); Acosta, L. [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); Amorini, F. [INFN - Laboratori Nazionali del Sud, Via S. Sofia, Catania (Italy); Auditore, L. [INFN Gruppo collegato di Messina and Dip. di Fisica e Scienze della Terra, Università di Messina (Italy); Berceanu, I. [Institute for Physics and Nuclear Engineering, Bucharest (Romania); Castoldi, A. [INFN Sezione di Milano e Politecnico Milano (Italy); De Filippo, E. [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); Dell' Aquila, D. [Dipartimento di scienze Fisiche, Università Federico II and INFN Sezione di Napoli (Italy); Francalanza, L. [INFN - Laboratori Nazionali del Sud, Via S. Sofia, Catania (Italy); Dip. di Fisica e Astronomia, Università di Catania, Via S. Sofia, Catania (Italy); Gnoffo, B. [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); Guazzoni, C. [INFN Sezione di Milano e Politecnico Milano (Italy); Lanzalone, G. [INFN - Laboratori Nazionali del Sud, Via S. Sofia, Catania (Italy); Facoltà di Ingegneria e Architettura, Università Kore, Enna (Italy); Lombardo, I. [Dipartimento di scienze Fisiche, Università Federico II and INFN Sezione di Napoli (Italy); Minniti, T.; Morgana, E.; Norella, S. [INFN Gruppo collegato di Messina and Dip. di Fisica e Scienze della Terra, Università di Messina (Italy); Pagano, A. [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); Pagano, E.V. [INFN - Laboratori Nazionali del Sud, Via S. Sofia, Catania (Italy); Dip. di Fisica e Astronomia, Università di Catania, Via S. Sofia, Catania (Italy); Papa, M.; Pirrone, S. [INFN - Sezione di Catania, Via S. Sofia, 95123 Catania (Italy); and others

    2015-11-01

    The gamma decay of the first excited 4.44 MeV 2+level of {sup 12}C, populated by inelastic scattering of proton and {sup 16}O beams at various energies was studied in order to test γ-ray detection efficiency and the quality of angular distribution information given by the CsI(Tl) detectors of the 4π CHIMERA array. The γ-decay was measured in coincidence with ejectile scattered particles in an approximately 4π geometry allowing to extract the angular distribution in the reference frame of recoiling {sup 12}C target. The typical sin{sup 2} (2θ) behavior of angular distribution was observed in the case of {sup 16}O beam. Besides that, for the proton beam, in order to explain the observed distribution, the addition of an incoherent flat contribution was required. This latter is the effect of proton spin flip events allowing the population of M=±1 magnetic substates, that is not possible in reactions induced by {sup 16}O beam. A comparison with previously collected data, obtained measuring only in and out of plane proton-γ-ray coincidences, confirms the good quality of the angular distribution information given by the apparatus. Possible applications with radioactive beams are outlined.

  5. Individual energy savings for individual flats in blocks of flats

    DEFF Research Database (Denmark)

    Nielsen, Anker; Rose, Jørgen

    2014-01-01

    and 1980. Normally, we expect the reduction in energy consumption to be around 20% for a 2 °C lower temperature, but for an inner flat the reduction can be up to 71%. The owners of the adjoining flats get an increase in energy demand of 10 to 20% each. They will not be able to figure out whether...... this is because the neighbour maintains a low temperature or the fact that they maintain a higher temperature. The best solution is to keep your own indoor temperature low. We can also turn the problem around: if you maintain a higher temperature than your neighbours, then you will pay part of their heating bill....

  6. Flat-on ambipolar triphenylamine/C60 nano-stacks formed from the self-organization of a pyramid-sphere-shaped amphiphile.

    Science.gov (United States)

    Liang, Wei-Wei; Huang, Chi-Feng; Wu, Kuan-Yi; Wu, San-Lien; Chang, Shu-Ting; Cheng, Yen-Ju; Wang, Chien-Lung

    2016-04-21

    A giant amphiphile, which is constructed with an amorphous nano-pyramid (triphenylamine, TPA) and a crystalline nano-sphere (C 60 ), was synthesized. Structural characterization indicates that this pyramid-sphere-shaped amphiphile ( TPA-C 60 ) forms a solvent-induced ordered phase, in which the two constituent units self-assemble into alternating stacks of two-dimensional (2D) TPA and C 60 nano-sheets. Due to the complexity of the molecular structure and the amorphous nature of the nano-pyramid, phase formation was driven by intermolecular C 60 -C 60 interactions and the ordered phase could not be reformed from the TPA-C 60 melt. Oriented crystal arrays of TPA-C 60 , which contain flat-on TPA/C 60 nano-stacks, can be obtained via a PDMS-assisted crystallization (PAC) technique. The flat-on dual-channel supramolecular structure of TPA-C 60 delivered ambipolar and balanced charge-transport characteristics with an average μ e of 2.11 × 10 -4 cm 2 V -1 s -1 and μ h of 3.37 × 10 -4 cm 2 V -1 s -1 . The anisotropic charge-transport ability of the pyramid-sphere-shaped amphiphile was further understood based on the lattice structure and the lattice orientation of TPA-C 60 revealed from electron diffraction analyses.

  7. Arteries of the falciform ligament on C-arm CT hepatic arteriography: The hepatic falciform artery and the Sappey's superior artery

    Energy Technology Data Exchange (ETDEWEB)

    Hur, Saebeom; Chung, Jin Wook; Lee, Jae Hwan; Cho, SooBeum; Kim, Minuk; Lee, Myungsu; Kim, Hyo-Cheol; Jae, Hwan Jun [Seoul National University Hospital, Department of Radiology, Seoul (Korea, Republic of); Zhou, Chun Gao [First Affiliated Hospital of Nanjing Medical University, Department of Interventional Radiology, Nanjing, Jangsu (China)

    2017-04-15

    To investigate the prevalence, anatomy and distribution of the hepatic falciform artery (HFA) and Sappey's superior artery (SSA) using C-arm CT hepatic arteriography (C-arm CTHA). From January 2011 to December 2012, 220 patients who underwent C-arm CTHA during initial transarterial treatment for hepatocellular carcinoma were included in this retrospective study. The HFAs and SSAs prevalence and origin were evaluated using axial images of C-arm CTHA. A 5-point scale for HFAs and a 4-point scale for SSAs were used to designate the radiologically conspicuous arteries. The prevalences of the total HFAs and SSAs were 95 % (n=209) and 22 % (n=49), while those of radiologically conspicuous HFAs and SSAs were 62 % (n=137) and 10 % (n=22), respectively. Thirty HFAs (22 % of radiologically conspicuous HFAs and 14 % of the total study population) were distributed in the subcutaneous layer of the anterior abdominal wall, while the majority of SSAs ran through the superior part of the falciform ligament in the left-anterior direction and anastomosed with left inferior phrenic artery. Our study using C-arm CTHA revealed that the prevalence of the HFA is higher than the existing knowledge and proved the existence of the SSA radiologically for the first time. (orig.)

  8. Arteries of the falciform ligament on C-arm CT hepatic arteriography: The hepatic falciform artery and the Sappey's superior artery

    International Nuclear Information System (INIS)

    Hur, Saebeom; Chung, Jin Wook; Lee, Jae Hwan; Cho, SooBeum; Kim, Minuk; Lee, Myungsu; Kim, Hyo-Cheol; Jae, Hwan Jun; Zhou, Chun Gao

    2017-01-01

    To investigate the prevalence, anatomy and distribution of the hepatic falciform artery (HFA) and Sappey's superior artery (SSA) using C-arm CT hepatic arteriography (C-arm CTHA). From January 2011 to December 2012, 220 patients who underwent C-arm CTHA during initial transarterial treatment for hepatocellular carcinoma were included in this retrospective study. The HFAs and SSAs prevalence and origin were evaluated using axial images of C-arm CTHA. A 5-point scale for HFAs and a 4-point scale for SSAs were used to designate the radiologically conspicuous arteries. The prevalences of the total HFAs and SSAs were 95 % (n=209) and 22 % (n=49), while those of radiologically conspicuous HFAs and SSAs were 62 % (n=137) and 10 % (n=22), respectively. Thirty HFAs (22 % of radiologically conspicuous HFAs and 14 % of the total study population) were distributed in the subcutaneous layer of the anterior abdominal wall, while the majority of SSAs ran through the superior part of the falciform ligament in the left-anterior direction and anastomosed with left inferior phrenic artery. Our study using C-arm CTHA revealed that the prevalence of the HFA is higher than the existing knowledge and proved the existence of the SSA radiologically for the first time. (orig.)

  9. Use of the mini C-arm for wrist fractures - Establishing a diagnostic reference level

    International Nuclear Information System (INIS)

    Love, G. J.; Pillai, A.; Gibson, S.

    2008-01-01

    The establishment of diagnostic reference levels (DRLs) for all typical radiological examinations became mandatory following the implementation of the Ionising Radiations (Medical Exposure) Regulations Act 2000. At present, there are no national dosage guidelines in the UK regarding use of fluoroscopy in orthopaedic trauma. The increasing popularity of the mini C-arm image intensifier amongst surgeons has led to concerns regarding use of ionizing radiation by personnel who have not been trained in radiation protection. It is therefore essential to have formal protocols for use of the mini C-arm to comply with the law and to maintain safe clinical practice. It is attempted to provide dose data for wrist fracture manipulations that may be used as a basis for setting a DRL for this procedure. Screening times were recorded for 80 wrist manipulations in a fracture clinic setting using a mini C-arm image intensifier. A DRL was set using the third quartile value for screening time. The median screening time for wrist fractures was 20 s with a range from 1 to 177 s. The third quartile value for screening time was 34 s. This value can be used as a provisional DRL for wrist fracture manipulations. The DRL is a quantitative guide for the optimisation of radiological protection. IR(ME)R 2000 states that if it is consistently exceeded by an individual operator or a piece of equipment, investigation and remedial action must be taken. We recommend that trauma units establish their own local DRLs for common procedures as made mandatory by legislation. (authors)

  10. Chlamydomonas DYX1C1/PF23 is essential for axonemal assembly and proper morphology of inner dynein arms.

    Directory of Open Access Journals (Sweden)

    Ryosuke Yamamoto

    2017-09-01

    Full Text Available Cytoplasmic assembly of ciliary dyneins, a process known as preassembly, requires numerous non-dynein proteins, but the identities and functions of these proteins are not fully elucidated. Here, we show that the classical Chlamydomonas motility mutant pf23 is defective in the Chlamydomonas homolog of DYX1C1. The pf23 mutant has a 494 bp deletion in the DYX1C1 gene and expresses a shorter DYX1C1 protein in the cytoplasm. Structural analyses, using cryo-ET, reveal that pf23 axonemes lack most of the inner dynein arms. Spectral counting confirms that DYX1C1 is essential for the assembly of the majority of ciliary inner dynein arms (IDA as well as a fraction of the outer dynein arms (ODA. A C-terminal truncation of DYX1C1 shows a reduction in a subset of these ciliary IDAs. Sucrose gradients of cytoplasmic extracts show that preassembled ciliary dyneins are reduced compared to wild-type, which suggests an important role in dynein complex stability. The role of PF23/DYX1C1 remains unknown, but we suggest that DYX1C1 could provide a scaffold for macromolecular assembly.

  11. Choosing a Motion Detector.

    Science.gov (United States)

    Ballard, David M.

    1990-01-01

    Examines the characteristics of three types of motion detectors: Doppler radar, infrared, and ultrasonic wave, and how they are used on school buses to prevent students from being killed by their own school bus. Other safety devices cited are bus crossing arms and a camera monitor system. (MLF)

  12. Real-Time Verification of a High-Dose-Rate Iridium 192 Source Position Using a Modified C-Arm Fluoroscope

    Energy Technology Data Exchange (ETDEWEB)

    Nose, Takayuki, E-mail: nose-takayuki@nms.ac.jp [Department of Radiation Oncology, Nippon Medical School Tamanagayama Hospital, Tama (Japan); Chatani, Masashi [Department of Radiation Oncology, Osaka Rosai Hospital, Sakai (Japan); Otani, Yuki [Department of Radiology, Kaizuka City Hospital, Kaizuka (Japan); Teshima, Teruki [Department of Radiation Oncology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka (Japan); Kumita, Shinichirou [Department of Radiology, Nippon Medical School Hospital, Tokyo (Japan)

    2017-03-15

    Purpose: High-dose-rate (HDR) brachytherapy misdeliveries can occur at any institution, and they can cause disastrous results. Even a patient's death has been reported. Misdeliveries could be avoided with real-time verification methods. In 1996, we developed a modified C-arm fluoroscopic verification of an HDR Iridium 192 source position prevent these misdeliveries. This method provided excellent image quality sufficient to detect errors, and it has been in clinical use at our institutions for 20 years. The purpose of the current study is to introduce the mechanisms and validity of our straightforward C-arm fluoroscopic verification method. Methods and Materials: Conventional X-ray fluoroscopic images are degraded by spurious signals and quantum noise from Iridium 192 photons, which make source verification impractical. To improve image quality, we quadrupled the C-arm fluoroscopic X-ray dose per pulse. The pulse rate was reduced by a factor of 4 to keep the average exposure compliant with Japanese medical regulations. The images were then displayed with quarter-frame rates. Results: Sufficient quality was obtained to enable observation of the source position relative to both the applicators and the anatomy. With this method, 2 errors were detected among 2031 treatment sessions for 370 patients within a 6-year period. Conclusions: With the use of a modified C-arm fluoroscopic verification method, treatment errors that were otherwise overlooked were detected in real time. This method should be given consideration for widespread use.

  13. Peginterferon alfa-2b and weight-based or flat-dose ribavirin in chronic hepatitis C patients: a randomized trial.

    Science.gov (United States)

    Jacobson, Ira M; Brown, Robert S; Freilich, Bradley; Afdhal, Nezam; Kwo, Paul Y; Santoro, John; Becker, Scott; Wakil, Adil E; Pound, David; Godofsky, Eliot; Strauss, Robert; Bernstein, David; Flamm, Steven; Pauly, Mary Pat; Mukhopadhyay, Pabak; Griffel, Louis H; Brass, Clifford A

    2007-10-01

    This prospective, multicenter, community-based and academic-based, open-label, investigator-initiated, U.S. study evaluated efficacy and safety of pegylated interferon (PEG-IFN) alfa-2b plus a flat or weight-based dose of ribavirin (RBV) in adults with chronic hepatitis C. Patients (n = 5027) were randomly assigned to receive PEG-IFN alfa-2b 1.5 microg/kg/week plus flat-dose (800 mg/day) or weight-based (800-1400 mg/day) RBV for 48 weeks (patients with genotype 1, 4, 5, or 6) and for 24 or 48 weeks (genotype 2/3 patients). Primary end point was sustained virologic response (undetectable [<125 IU/mL] serum hepatitis C virus RNA at 24-week follow-up). Sustained virologic response, but not end-of-treatment, rates were significantly higher with weight-based than with flat-dose RBV (44.2% versus 40.5%; P = 0.008). Sustained virologic response rates by intention-to-treat analysis were 34.0% and 28.9%, respectively, in genotype 1 patients (P = 0.005) and 31.2% and 26.7%, respectively, in genotype 1 patients with high baseline viral load (P = 0.056). In genotype 2/3 patients, rates were not significantly different (61.8% and 59.5%, respectively) regardless of treatment duration. Besides greater hemoglobin reductions with weight-based RBV, safety profiles were similar across RBV dosing groups, including the 1400-mg/day group. PEG-IFN alfa-2b plus weight-based RBV is more effective than flat-dose RBV, particularly in genotype 1 patients, providing equivalent efficacy across all weight groups. RBV 1400 mg/day is appropriate for patients 105 to 125 kg. For genotype 2/3 patients, 24 weeks of treatment with flat-dose RBV is adequate; no evidence of additional benefit of extending treatment to 48 weeks was demonstrated.

  14. Performance of low-cost X-ray area detectors with consumer digital cameras

    International Nuclear Information System (INIS)

    Panna, A.; Gomella, A.A.; Harmon, K.J.; Chen, P.; Miao, H.; Bennett, E.E.; Wen, H.

    2015-01-01

    We constructed X-ray detectors using consumer-grade digital cameras coupled to commercial X-ray phosphors. Several detector configurations were tested against the Varian PaxScan 3024M (Varian 3024M) digital flat panel detector. These include consumer cameras (Nikon D800, Nikon D700, and Nikon D3X) coupled to a green emission phosphor in a back-lit, normal incidence geometry, and in a front-lit, oblique incidence geometry. We used the photon transfer method to evaluate detector sensitivity and dark noise, and the edge test method to evaluate their spatial resolution. The essential specifications provided by our evaluation include discrete charge events captured per mm 2 per unit exposure surface dose, dark noise in equivalents of charge events per pixel, and spatial resolution in terms of the full width at half maximum (FWHM) of the detector's line spread function (LSF). Measurements were performed using a tungsten anode X-ray tube at 50 kVp. The results show that the home-built detectors provide better sensitivity and lower noise than the commercial flat panel detector, and some have better spatial resolution. The trade-off is substantially smaller imaging areas. Given their much lower costs, these home-built detectors are attractive options for prototype development of low-dose imaging applications

  15. Development of a fast multi-line x-ray CT detector for NDT

    International Nuclear Information System (INIS)

    Hofmann, T.; Nachtrab, F.; Schlechter, T.; Mühlbauer, J.; Schröpfer, S.; Firsching, M.; Uhlmann, N.; Neubauer, H.; Ernst, J.; Schweiger, T.; Oberst, M.; Meyer, A.

    2015-01-01

    Typical X-ray detectors for non-destructive testing (NDT) are line detectors or area detectors, like e.g. flat panel detectors. Multi-line detectors are currently only available in medical Computed Tomography (CT) scanners. Compared to flat panel detectors, line and multi-line detectors can achieve much higher frame rates. This allows time-resolved 3D CT scans of an object under investigation. Also, an improved image quality can be achieved due to reduced scattered radiation from object and detector themselves. Another benefit of line and multi-line detectors is that very wide detectors can be assembled easily, while flat panel detectors are usually limited to an imaging field with a size of approx. 40 × 40 cm 2 at maximum. The big disadvantage of line detectors is the limited number of object slices that can be scanned simultaneously. This leads to long scan times for large objects. Volume scans with a multi-line detector are much faster, but with almost similar image quality. Due to the promising properties of multi-line detectors their application outside of medical CT would also be very interesting for NDT. However, medical CT multi-line detectors are optimized for the scanning of human bodies. Many non-medical applications require higher spatial resolutions and/or higher X-ray energies. For those non-medical applications we are developing a fast multi-line X-ray detector.In the scope of this work, we present the current state of the development of the novel detector, which includes several outstanding properties like an adjustable curved design for variable focus-detector-distances, conserving nearly uniform perpendicular irradiation over the entire detector width. Basis of the detector is a specifically designed, radiation hard CMOS imaging sensor with a pixel pitch of 200 μ m. Each pixel has an automatic in-pixel gain adjustment, which allows for both: a very high sensitivity and a wide dynamic range. The final detector is planned to have 256 lines of

  16. Musculoskeletal imaging with a prototype photon-counting detector.

    Science.gov (United States)

    Gruber, M; Homolka, P; Chmeissani, M; Uffmann, M; Pretterklieber, M; Kainberger, F

    2012-01-01

    To test a digital imaging X-ray device based on the direct capture of X-ray photons with pixel detectors, which are coupled with photon-counting readout electronics. The chip consists of a matrix of 256 × 256 pixels with a pixel pitch of 55 μm. A monolithic image of 11.2 cm × 7 cm was obtained by the consecutive displacement approach. Images of embalmed anatomical specimens of eight human hands were obtained at four different dose levels (skin dose 2.4, 6, 12, 25 μGy) with the new detector, as well as with a flat-panel detector. The overall rating scores for the evaluated anatomical regions ranged from 5.23 at the lowest dose level, 6.32 at approximately 6 μGy, 6.70 at 12 μGy, to 6.99 at the highest dose level with the photon-counting system. The corresponding rating scores for the flat-panel detector were 3.84, 5.39, 6.64, and 7.34. When images obtained at the same dose were compared, the new system outperformed the conventional DR system at the two lowest dose levels. At the higher dose levels, there were no significant differences between the two systems. The photon-counting detector has great potential to obtain musculoskeletal images of excellent quality at very low dose levels.

  17. Transport processes in intertidal sand flats

    Science.gov (United States)

    Wu, Christy

    2010-05-01

    Methane rich sulfate depleted seeps are observed along the low water line of the intertidal sand flat Janssand in the Wadden Sea. It is unclear where in the flat the methane is formed, and how it is transported to the edge of the sand flat where the sulfidic water seeps out. Methane and sulfate distributions in pore water were determined along transects from low water line toward the central area of the sand flat. The resulting profiles showed a zone of methane-rich and sulfate-depleted pore water below 2 m sediment depth. Methane production and sulfate reduction are monitored over time for surface sediments collected from the upper flat and seeping area. Both activities were at 22 C twice as high as at 15 C. The rates in sediments from the central area were higher than in sediments from the methane seeps. Methanogenesis occurred in the presence of sulfate, and was not significantly accelerated when sulfate was depleted. The observations show a rapid anaerobic degradation of organic matter in the Janssand. The methane rich pore water is obviously transported with a unidirectional flow from the central area of the intertidal sand flat toward the low water line. This pore water flow is driven by the pressure head caused by elevation of the pore water relative to the sea surface at low tide (Billerbeck et al. 2006a). The high methane concentration at the low water line accumulates due to a continuous outflow of pore water at the seepage site that prevents penetration of electron acceptors such as oxygen and sulfate to reoxidize the reduced products of anaerobic degradation (de Beer et al. 2006). It is, however, not clear why no methane accumulates or sulfate is depleted in the upper 2 m of the flats.

  18. Concept of data storage prototype for Super-C-Tau factory detector

    International Nuclear Information System (INIS)

    Maximov, D.A.

    2017-01-01

    The physics program of experiments at the Super- c τ factory with a peak luminosity of 10 35 cm −2 s −1 leads to high requrements for Data Acquisition and Data Storage systems. Detector data storage is one of the key component of the detector infrastructure, so it must be reliable, highly available and fault tolerant shared storage. It is mostly oriented (from end user point of view) for sequential but mixed read and write operations and is planed to store large data blocks (files). According to CDR of Super-C-Tau factory detector data storage must have very high performance (up to 1 Tbps in both directions simultaneously) and have significant volume (tens and hundreds of Petabytes). It is decided to build a series of prototypes with growing capabilities to investigate storage and neighboring technologies. First prototype of data storage is aimed to develop and test basic components of detector data storage system such as storage devices, networks and software. This prototype is designed to be capable to work with data rate of order 10 Gbps. It is estimated that about 5 modern computers with about 50 disks in total should be enough to archive required performance. The prototype will be based on Ceph storage technology. Ceph is a distributed storage system which allows to create storage solutions with very flexible design, high availability and scalability.

  19. Scan equalization digital radiography (SEDR) implemented with an amorphous selenium flat-panel detector: initial experience

    Energy Technology Data Exchange (ETDEWEB)

    Liu Xinming; Lai Chaojen; Chen Lingyun; Han Tao; Zhong Yuncheng; Shen Youtao; Wang Tianpeng; Shaw, Chris C [Department of Imaging Physics, Digital Imaging Research Laboratory, University of Texas M. D. Anderson Cancer Center, Houston, TX 77030-4009 (United States)], E-mail: xliu@di.mdacc.tmc.edu

    2009-11-21

    It is well recognized in projection radiography that low-contrast detectability suffers in heavily attenuating regions due to excessively low x-ray fluence to the image receptor and higher noise levels. Exposure equalization can improve image quality by increasing the x-ray exposure to heavily attenuating regions, resulting in a more uniform distribution of exposure to the detector. Image quality is also expected to be improved by using the slot-scan geometry to reject scattered radiation effectively without degrading primary x-rays. This paper describes the design of a prototype scan equalization digital radiography (SEDR) system implemented with an amorphous silicon (a-Si) thin-film transistor (TFT) array-based flat-panel detector. With this system, slot-scan geometry with alternate line erasure and readout (ALER) technique was used to achieve scatter rejection. A seven-segment beam height modulator assembly was mounted onto the fore collimator to regulate exposure regionally for chest radiography. The beam modulator assembly, consisting of micro linear motors, lead screw cartridge with lead (Pb) beam blockers attached, position feedback sensors and motor driver circuitry, has been tested and found to have an acceptable response for exposure equalization in chest radiography. An anthropomorphic chest phantom was imaged in the posterior-anterior (PA) view under clinical conditions. Scatter component, primary x-rays, scatter-to-primary ratios (SPRs) and primary signal-to-noise ratios (PSNRs) were measured in the SEDR images to evaluate the rejection and redistribution of scattered radiation, and compared with those for conventional full-field imaging with and without anti-scatter grid methods. SPR reduction ratios (SPRRRs, defined as the differences between the non-grid full-field SPRs and the reduced SPRs divided by the former) yielded approximately 59% for the full-field imaging with grid and 82% for the SEDR technique in the lungs, and 77% for the full

  20. Prospects and Results from the AFP Detector in ATLAS

    CERN Document Server

    Gach, Grzegorz; The ATLAS collaboration

    2016-01-01

    Status of the AFP project in the ATLAS experiment is given. In 2016 one arm of the AFP detector was installed and first data have been taken. In parallel with integration of the AFP subdetector into the ATLAS TDAQ nad DCS, beam tests and preparations for the installation of the 2nd arm are performed.

  1. Utility of C-arm CT in overcoming challenges in patients undergoing Transarterial chemoembolization for hepatocellular carcinoma

    International Nuclear Information System (INIS)

    Kulkarni, Chinmay; Sreekumar, K. P.; Prabhu, Nirmal Kumar; Kannan, Rajesh R; Moorthy, Srikanth

    2014-01-01

    Transarterial chemoembolization (TACE) is the well-known treatment for hepatocellular carcinoma. Multiple digital subtraction angiography (DSA) acquisitions in different projections are required to identify difficult arterial feeders. Moreover, the tell-tale tumor blush can be obscured by proximity to lung base, small size of lesion, and breathing artifacts. C-arm CT is a revolutionary advancement in the intervention radiology suite that allows acquisition of data which can be reformatted in multiple planes and volume rendered incorporating both soft tissue and vascular information like multidetector computed tomography (MDCT). These images acquired during the TACE procedure can provide critical inputs for achieving a safe and effective therapy. This case series aims to illustrate the utility of C-arm CT in solving specific problems encountered while performing TACE

  2. SU-E-CAMPUS-I-04: Automatic Skin-Dose Mapping for An Angiographic System with a Region-Of-Interest, High-Resolution Detector

    Energy Technology Data Exchange (ETDEWEB)

    Vijayan, S; Rana, V [Department of Physiology and Biophysics, Toshiba Stroke and Vascular Research Center (United States); Setlur Nagesh, S [Toshiba Stroke and Vascular Research Center (United States); Ionita, C [Department of Biomedical Engineering, University at Buffalo (State University of New York), Buffalo, NY (United States); Rudin, S [Department of Radiology, Department of Physiology and Biophysics, Toshiba Stroke and Vascular Research Center, Department of Biomedical Engineering, University at Buffalo (State University of New York), Buffalo, NY (United States); Bednarek, D [Department of Radiology, Department of Physiology and Biophysics, Toshiba Stroke and Vascular Research Center (United States)

    2014-06-15

    Purpose: Our real-time skin dose tracking system (DTS) has been upgraded to monitor dose for the micro-angiographic fluoroscope (MAF), a high-resolution, small field-of-view x-ray detector. Methods: The MAF has been mounted on a changer on a clinical C-Arm gantry so it can be used interchangeably with the standard flat-panel detector (FPD) during neuro-interventional procedures when high resolution is needed in a region-of-interest. To monitor patient skin dose when using the MAF, our DTS has been modified to automatically account for the change in scatter for the very small MAF FOV and to provide separated dose distributions for each detector. The DTS is able to provide a color-coded mapping of the cumulative skin dose on a 3D graphic model of the patient. To determine the correct entrance skin exposure to be applied by the DTS, a correction factor was determined by measuring the exposure at the entrance surface of a skull phantom with an ionization chamber as a function of entrance beam size for various beam filters and kVps. Entrance exposure measurements included primary radiation, patient backscatter and table forward scatter. To allow separation of the dose from each detector, a parameter log is kept that allows a replay of the procedure exposure events and recalculation of the dose components.The graphic display can then be constructed showing the dose distribution from the MAF and FPD separately or together. Results: The DTS is able to provide separate displays of dose for the MAF and FPD with field-size specific scatter corrections. These measured corrections change from about 49% down to 10% when changing from the FPD to the MAF. Conclusion: The upgraded DTS allows identification of the patient skin dose delivered when using each detector in order to achieve improved dose management as well as to facilitate peak skin-dose reduction through dose spreading. Research supported in part by Toshiba Medical Systems Corporation and NIH Grants R43FD0158401, R44FD

  3. Standard practice for radiologic examination of flat panel composites and sandwich core materials used in aerospace applications

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This practice is intended to be used as a supplement to Practices E 1742, E 1255, and E 2033. 1.2 This practice describes procedures for radiologic examination of flat panel composites and sandwich core materials made entirely or in part from fiber-reinforced polymer matrix composites. Radiologic examination is: a) radiographic (RT) with film, b) Computed Radiography (CR) with Imaging Plate, c) Digital Radiology (DR) with Digital Detector Array’s (DDA), and d) Radioscopic (RTR) Real Time Radiology with a detection system such as an Image Intensifier. The composite materials under consideration typically contain continuous high modulus fibers (> 20 GPa), such as those listed in 1.4. 1.3 This practice describes established radiological examination methods that are currently used by industry that have demonstrated utility in quality assurance of flat panel composites and sandwich core materials during product process design and optimization, process control, after manufacture inspection, in service exami...

  4. Flat H Redundant Frangible Joint Development

    Science.gov (United States)

    Brown, Chris

    2016-01-01

    Orion and Commercial Crew Program (CCP) Partners have chosen to use frangible joints for certain separation events. The joints currently available are zero failure tolerant and will be used in mission safety applications. The goal is to further develop a NASA designed redundant frangible joint that will lower flight risk and increase reliability. FY16 testing revealed a successful design in subscale straight test specimens that gained efficiency and supports Orion load requirements. Approach / Innovation A design constraint is that the redundant joint must fit within the current Orion architecture, without the need for additional vehicle modification. This limitation required a design that changed the orientation of the expanding tube assemblies (XTAs), by rotating them 90deg from the standard joint configuration. The change is not trivial and affects the fracture mechanism and structural load paths. To address these changes, the design incorporates cantilevered arms on the break plate. The shock transmission and expansion of the XTA applies force to these arms and creates a prying motion to push the plate walls outward to the point of structural failure at the notched section. The 2014 test design revealed that parts could slip during functioning wasting valuable energy needed to separate the structure with only a single XTA functioning. Dual XTA functioning fully separated the assembly showing a discrepancy can be backed up with redundancy. Work on other fully redundant systems outside NASA is limited to a few patents that have not been subjected to functionality testing Design changes to prevent unwanted slippage (with ICA funding in 2015) showed success with a single XTA. The main goal for FY 2016 was to send the new Flat H RFJ to WSTF where single XTA test failures occurred back in 2014. The plan was to gain efficiency in this design by separating the Flat H RFJ with thicker ligaments with dimensions baselined in 2014. Other modifications included geometry

  5. Detector development and experiments at COSY

    International Nuclear Information System (INIS)

    Morsch, H.P.

    1988-05-01

    These proceedings contain the manuscripts of the lectures presented at the named workshop. These concern a review about the COSY project, ideal detectors for hadron physics at COSY, possible experiments at COSY, magnetic spectrometers, a modification of BIG KARL, consideration on COSY experiments in the early stage, a detector for exclusive 2-meson production experiments, the excitation of baryons and physics with complex projectiles, a status report about the Indiana cooler ring, special scintillators, multiwire chambers, position-sensitive semiconductor detectors, detectors for neutral particles, a small large-acceptance photon detector, a status report of the two-arm photon spectrometer TAPS, studies on the parity violation in the pp scattering, the measurement of excitation functions for the study of dibaryon states, and results from the neutron workshop held in February 1988 at the KFA Juelich. (HSI)

  6. Pentachlorophenol induction of the Pseudomonas aeruginosa mexAB-oprM efflux operon: involvement of repressors NalC and MexR and the antirepressor ArmR.

    Directory of Open Access Journals (Sweden)

    Lisa M Starr

    Full Text Available Pentachlorophenol (PCP induced expression of the NalC repressor-regulated PA3720-armR operon and the MexR repressor-controlled mexAB-oprM multidrug efflux operon of Pseudomonas aeruginosa. PCP's induction of PA3720-armR resulted from its direct modulation of NalC, the repressor's binding to PA3720-armR promoter-containing DNA as seen in electromobility shift assays (EMSAs being obviated in the presence of this agent. The NalC binding site was localized to an inverted repeat (IR sequence upstream of PA3720-armR and overlapping a promoter region whose transcription start site was mapped. While modulation of MexR by the ArmR anti-repressor explains the upregulation of mexAB-oprM in nalC mutants hyperexpressing PA3720-armR, the induction of mexAB-oprM expression by PCP is not wholly explainable by PCP induction of PA3720-armR and subsequent ArmR modulation of MexR, inasmuch as armR deletion mutants still showed PCP-inducible mexAB-oprM expression. PCP failed, however, to induce mexAB-oprM in a mexR deletion strain, indicating that MexR was required for this, although PCP did not modulate MexR binding to mexAB-oprM promoter-containing DNA in vitro. One possibility is that MexR responds to PCP-generated in vivo effector molecules in controlling mexAB-oprM expression in response to PCP. PCP is an unlikely effector and substrate for NalC and MexAB-OprM--its impact on NalC binding to the PA3720-armR promoter DNA occurred only at high µM levels--suggesting that it mimics an intended phenolic effector/substrate(s. In this regard, plants are an abundant source of phenolic antimicrobial compounds and, so, MexAB-OprM may function to protect P. aeruginosa from plant antimicrobials that it encounters in nature.

  7. Plasma-laser ion discrimination by TOF technique applied to coupled SiC detectors.

    Science.gov (United States)

    Cavallaro, Salvatore

    2018-01-01

    The rate estimation of nuclear reactions induced in high intensity laser-target interaction (≥1016 W/cm2), is strongly depending on the neutron detection efficiency and ion charge discrimination, according to particles involved in exit open-channels. Ion discrimination is basically performed by means of analysis of pits observed on track detector, which is critically dependent on calibration and/or fast TOF devices based on SiC and diamond detectors. Last setup is used to determine the ion energy and to obtain a rough estimation of yields. However, for each TOF interval, the dependence of yield from the energy deposited in the detector sensitive region, introduces a distortion in the ion spectra. Moreover, if two ion species are present in the same spectrum, the discrimination of their contribution is not attainable. In this paper a new method is described which allows to discriminate the contribution of two ion species in the wide energy range of nuclear reactions induced in laser-target interactions. The method is based on charge response of two TOF-SiC detectors, of suitable thicknesses, placed in adjacent positions. In presence of two ion species, the response of the detectors, associated with different energy losses, can determine the ion specific contribution to each TOF interval.

  8. Interventional heart wall motion analysis with cardiac C-arm CT systems

    International Nuclear Information System (INIS)

    Müller, Kerstin; Maier, Andreas K; Schwemmer, Chris; Hornegger, Joachim; Zheng, Yefeng; Wang, Yang; Lauritsch, Günter; Rohkohl, Christopher; Fahrig, Rebecca

    2014-01-01

    Today, quantitative analysis of three-dimensional (3D) dynamics of the left ventricle (LV) cannot be performed directly in the catheter lab using a current angiographic C-arm system, which is the workhorse imaging modality for cardiac interventions. Therefore, myocardial wall analysis is completely based on the 2D angiographic images or pre-interventional 3D/4D imaging. In this paper, we present a complete framework to study the ventricular wall motion in 4D (3D+t) directly in the catheter lab. From the acquired 2D projection images, a dynamic 3D surface model of the LV is generated, which is then used to detect ventricular dyssynchrony. Different quantitative features to evaluate LV dynamics known from other modalities (ultrasound, magnetic resonance imaging) are transferred to the C-arm CT data. We use the ejection fraction, the systolic dyssynchrony index a 3D fractional shortening and the phase to maximal contraction (ϕ i, max ) to determine an indicator of LV dyssynchrony and to discriminate regionally pathological from normal myocardium. The proposed analysis tool was evaluated on simulated phantom LV data with and without pathological wall dysfunctions. The LV data used is publicly available online at https://conrad.stanford.edu/data/heart. In addition, the presented framework was tested on eight clinical patient data sets. The first clinical results demonstrate promising performance of the proposed analysis tool and encourage the application of the presented framework to a larger study in clinical practice. (paper)

  9. A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization

    Science.gov (United States)

    Meyer, Michael; Kalender, Willi A.; Kyriakou, Yiannis

    2010-01-01

    Scattered radiation is a major source of artifacts in flat detector computed tomography (FDCT) due to the increased irradiated volumes. We propose a fast projection-based algorithm for correction of scatter artifacts. The presented algorithm combines a convolution method to determine the spatial distribution of the scatter intensity distribution with an object-size-dependent scaling of the scatter intensity distributions using a priori information generated by Monte Carlo simulations. A projection-based (PBSE) and an image-based (IBSE) strategy for size estimation of the scanned object are presented. Both strategies provide good correction and comparable results; the faster PBSE strategy is recommended. Even with such a fast and simple algorithm that in the PBSE variant does not rely on reconstructed volumes or scatter measurements, it is possible to provide a reasonable scatter correction even for truncated scans. For both simulations and measurements, scatter artifacts were significantly reduced and the algorithm showed stable behavior in the z-direction. For simulated voxelized head, hip and thorax phantoms, a figure of merit Q of 0.82, 0.76 and 0.77 was reached, respectively (Q = 0 for uncorrected, Q = 1 for ideal). For a water phantom with 15 cm diameter, for example, a cupping reduction from 10.8% down to 2.1% was achieved. The performance of the correction method has limitations in the case of measurements using non-ideal detectors, intensity calibration, etc. An iterative approach to overcome most of these limitations was proposed. This approach is based on root finding of a cupping metric and may be useful for other scatter correction methods as well. By this optimization, cupping of the measured water phantom was further reduced down to 0.9%. The algorithm was evaluated on a commercial system including truncated and non-homogeneous clinically relevant objects.

  10. "Flat-Fish" Vacuum Chamber

    CERN Multimedia

    CERN PhotoLab

    1978-01-01

    The picture shows a "Flat-Fish" vacuum chamber being prepared in the ISR workshop for testing prior to installation in the Split Field Magnet (SFM) at intersection I4. The two shells of each part were hydroformed from 0.15 mm thick inconel 718 sheet (with end parts in inconel 600 for easier manual welding to the arms) and welded toghether with two strips which were attached by means of thin stainless steel sheets to the Split Field Magnet poles in order to take the vertical component of the atmospheric pressure force. This was the thinnest vacuum chamber ever made for the ISR. Inconel material was chosen for its high elastic modulus and strenght at chamber bake-out temperature. In this picture the thin sheets transferring the vertical component of the atmosferic pressure force are attached to a support frame for testing. See also 7712182, 7712179.

  11. THE B/C AND SUB-IRON/IRON COSMIC RAY RATIOS—FURTHER EVIDENCE IN FAVOR OF THE SPIRAL-ARM DIFFUSION MODEL

    International Nuclear Information System (INIS)

    Benyamin, David; Piran, Tsvi; Shaviv, Nir J.; Nakar, Ehud

    2016-01-01

    The boron to carbon (B/C) and sub-Fe/Fe ratios provide an important clue on cosmic ray (CR) propagation within the Galaxy. These ratios estimate the grammage that the CRs traverse as they propagate from their sources to Earth. Attempts to explain these ratios within the standard CR propagation models require ad hoc modifications and even with those these models necessitate inconsistent grammages to explain both ratios. As an alternative, physically motivated model, we have proposed that CRs originate preferably within the galactic spiral arms. CR propagation from dynamic spiral arms has important imprints on various secondary to primary ratios, such as the B/C ratio and the positron fraction. We use our spiral-arm diffusion model with the spallation network extended up to nickel to calculate the sub-Fe/Fe ratio. We show that without any additional parameters the spiral-arm model consistently explains both ratios with the same grammage, providing further evidence in favor of this model.

  12. THE B/C AND SUB-IRON/IRON COSMIC RAY RATIOS—FURTHER EVIDENCE IN FAVOR OF THE SPIRAL-ARM DIFFUSION MODEL

    Energy Technology Data Exchange (ETDEWEB)

    Benyamin, David; Piran, Tsvi; Shaviv, Nir J. [The Racah Institute of Physics, The Hebrew University of Jerusalem, Jerusalem 91904 (Israel); Nakar, Ehud [Raymond and Beverly Sackler School of Physics and Astronomy, Tel Aviv University, Tel Aviv 69978 (Israel)

    2016-07-20

    The boron to carbon (B/C) and sub-Fe/Fe ratios provide an important clue on cosmic ray (CR) propagation within the Galaxy. These ratios estimate the grammage that the CRs traverse as they propagate from their sources to Earth. Attempts to explain these ratios within the standard CR propagation models require ad hoc modifications and even with those these models necessitate inconsistent grammages to explain both ratios. As an alternative, physically motivated model, we have proposed that CRs originate preferably within the galactic spiral arms. CR propagation from dynamic spiral arms has important imprints on various secondary to primary ratios, such as the B/C ratio and the positron fraction. We use our spiral-arm diffusion model with the spallation network extended up to nickel to calculate the sub-Fe/Fe ratio. We show that without any additional parameters the spiral-arm model consistently explains both ratios with the same grammage, providing further evidence in favor of this model.

  13. Towards radiation hard converter material for SiC-based fast neutron detectors

    Science.gov (United States)

    Tripathi, S.; Upadhyay, C.; Nagaraj, C. P.; Venkatesan, A.; Devan, K.

    2018-05-01

    In the present work, Geant4 Monte-Carlo simulations have been carried out to study the neutron detection efficiency of the various neutron to other charge particle (recoil proton) converter materials. The converter material is placed over Silicon Carbide (SiC) in Fast Neutron detectors (FNDs) to achieve higher neutron detection efficiency as compared to bare SiC FNDs. Hydrogenous converter material such as High-Density Polyethylene (HDPE) is preferred over other converter materials due to the virtue of its high elastic scattering reaction cross-section for fast neutron detection at room temperature. Upon interaction with fast neutrons, hydrogenous converter material generates recoil protons which liberate e-hole pairs in the active region of SiC detector to provide a detector signal. The neutron detection efficiency offered by HDPE converter is compared with several other hydrogenous materials viz., 1) Lithium Hydride (LiH), 2) Perylene, 3) PTCDA . It is found that, HDPE, though providing highest efficiency among various studied materials, cannot withstand high temperature and harsh radiation environment. On the other hand, perylene and PTCDA can sustain harsh environments, but yields low efficiency. The analysis carried out reveals that LiH is a better material for neutron to other charge particle conversion with competent efficiency and desired radiation hardness. Further, the thickness of LiH has also been optimized for various mono-energetic neutron beams and Am-Be neutron source generating a neutron fluence of 109 neutrons/cm2. The optimized thickness of LiH converter for fast neutron detection is found to be ~ 500 μm. However, the estimated efficiency for fast neutron detection is only 0.1%, which is deemed to be inadequate for reliable detection of neutrons. A sensitivity study has also been done investigating the gamma background effect on the neutron detection efficiency for various energy threshold of Low-Level Discriminator (LLD). The detection

  14. ATLAS Forward Proton Detector

    CERN Document Server

    Grieco, Chiara; The ATLAS collaboration

    2018-01-01

    The aim of the ATLAS Forward Proton (AFP) detector system is the measurement of protons scattered diffractively or electromagnetically at very small angles. The full two-arm setup was installed during the 2016/2017 EYETS. This allows measurements of processes with two forward protons: central diffraction, exclusive production, and two-photon processes. In 2017, AFP participated in the ATLAS high-luminosity data taking on the day-by-day basis. In addition, several special runs with reduced luminosity were taken. The poster will present the AFP detectors and the lessons learned from the last year operation and some performance from 2016 and 2017.

  15. ALICE Transition Radiation Detector

    CERN Multimedia

    Pachmayer, Y

    2013-01-01

    The Transition Radiation Detector (TRD) is the main electron detector in ALICE. In conduction with the TPC and the ITS, it provides the necessary electron identification capability to study: - Production of light and heavy vector mesons as well as the continuum in the di-electron channel, - Semi leptonic decays of hadrons with open charm and open beauty via the single-electron channel using the displaced vertex information provided by the ITS, - Correlated DD and BB pairs via coincidences of electrons in the central barrel and muons in the forward muon arm, - Jets with high Pτ tracks in one single TRD stack.

  16. Analysis of Patients' X-ray Exposure in 146 Percutaneous Radiologic Gastrostomies.

    Science.gov (United States)

    Petersen, Tim-Ole; Reinhardt, Martin; Fuchs, Jochen; Gosch, Dieter; Surov, Alexey; Stumpp, Patrick; Kahn, Thomas; Moche, Michael

    2017-09-01

    significantly extends FT, but has little effect on the overall dose of the intervention. Due to the significantly higher X-ray exposure, the use of a CBCT as well as PRG-placements using CT should be limited to clinically absolutely necessary exceptions with strict indication. Key Points   · Fluoroscopically-guided PRG placements are interventions with low X-ray exposure.. · X-ray exposure from fluoroscopy is lower using flat panel detector systems as compared to image intensifier systems.. · The concomitant placement of an oro- or nasogastric probe extends the fluoroscopy time.. · Gastric probe placement is worthwhile to prevent the premature use of the significantly radiation-intensive CT.. · The use of the C-arm CT or the CT increases the beam exposure by 26 or 33 times, respectively.. · The PRG placement using C-arm CT and CT should only be performed in exceptional cases.. Citation Format · Petersen TO, Reinhardt M, Fuchs J et al. Analysis of Patients' X-ray Exposure in 146 Percutaneous Radiologic Gastrostomies. Fortschr Röntgenstr 2017; 189: 820 - 827. © Georg Thieme Verlag KG Stuttgart · New York.

  17. ARM Airborne Carbon Measurements VI (ARM-ACME VI) Field Campaign Report

    Energy Technology Data Exchange (ETDEWEB)

    Biraud, Sebastien [Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)

    2017-05-01

    From October 1, 2015 through September 30, 2016, AAF deployed a Cessna 206 aircraft over the Southern Great Plains, collecting observations of trace gas mixing ratios over the ARM/SGP Central Facility. The aircraft payload included two Atmospheric Observing Systems (AOS Inc.) analyzers for continuous measurements of CO2, and a 12-flask sampler for analysis of carbon cycle gases (CO2, CO, CH4, N2O, 13CO2). The aircraft payload also includes solar/infrared radiation measurements. This research (supported by DOE ARM and TES programs) builds upon previous ARM-ACME missions. The goal of these measurements is to improve understanding of: (a) the carbon exchange of the ARM region; (b) how CO2 and associated water and energy fluxes influence radiative forcing, convective processes, and CO2 concentrations over the ARM region, and (c) how greenhouse gases are transported on continental scales.

  18. Analysis of radiation risk to patients from intra-operative use of the mobile X-ray system (C-arm

    Directory of Open Access Journals (Sweden)

    Yang-Sub Lee

    2015-01-01

    Full Text Available Background: The aim of this study was to investigate clinical applications of mobile C-arms and consequent radiation risk, to increase medical attention on radiation protection, and to provide basic data for safe radiation use in the operating room. Materials and Methods: In this study, a total of 374 surgical operations, conducted using a portable fluoroscopic X-ray system from January to March of 2013, were analyzed. Dose summaries produced by the General Electric C-arm and data elements in digital imaging and communications in the medicine header of Ziehm C-arm, fluoroscopy time were used to obtain dose-area product (DAP and effective dose. Corresponding mean and maximum values were calculated, and the resulting data on the frequency of application, fluoroscopy time, DAP, and effective dose were compared and analyzed in terms of surgical specialty and operation types. Results: Orthopedic surgery was the most frequent with 165 cases (44.1%. The highest DAP value and effective dose were found in liver transplant among surgical specialty fields, with mean values of 2.90 ± 3.76 mGy∙m 2 and 58 ± 75.2 mSv, respectively (P = 0.0001. The highest DAP value and effective dose were observed in intra-operative mesenteric portography among types of surgery, showing mean values of 2.90 ± 3.81 mGy∙m 2 and 58.03 ± 76.24 mSv, respectively (P = 0.0001. Conclusion: Because DAP varies significantly across surgical specialties and types of operation, aggressive efforts to understand the effects of radiation dose is critical for radiation protection from intra-operative use of mobile C-arms.

  19. Prospects and Results from the AFP Detector in ATLAS

    CERN Document Server

    Gach, Grzegorz; The ATLAS collaboration

    2017-01-01

    In 2016 one arm of the AFP detector was installed and first data have been taken. In parallel with integration of the AFP subdetector into the ATLAS TDAQ and DCS systems, beam tests and preparations for the installation of the 2$^{\\textrm{nd}}$ arm are performed. In this report, a status of the AFP project in the ATLAS experiment is discussed.

  20. Flat tori in three-dimensional space and convex integration.

    Science.gov (United States)

    Borrelli, Vincent; Jabrane, Saïd; Lazarus, Francis; Thibert, Boris

    2012-05-08

    It is well-known that the curvature tensor is an isometric invariant of C(2) Riemannian manifolds. This invariant is at the origin of the rigidity observed in Riemannian geometry. In the mid 1950s, Nash amazed the world mathematical community by showing that this rigidity breaks down in regularity C(1). This unexpected flexibility has many paradoxical consequences, one of them is the existence of C(1) isometric embeddings of flat tori into Euclidean three-dimensional space. In the 1970s and 1980s, M. Gromov, revisiting Nash's results introduced convex integration theory offering a general framework to solve this type of geometric problems. In this research, we convert convex integration theory into an algorithm that produces isometric maps of flat tori. We provide an implementation of a convex integration process leading to images of an embedding of a flat torus. The resulting surface reveals a C(1) fractal structure: Although the tangent plane is defined everywhere, the normal vector exhibits a fractal behavior. Isometric embeddings of flat tori may thus appear as a geometric occurrence of a structure that is simultaneously C(1) and fractal. Beyond these results, our implementation demonstrates that convex integration, a theory still confined to specialists, can produce computationally tractable solutions of partial differential relations.

  1. Ultra low-dose VCUG in children using a modern flat detectorunit

    Energy Technology Data Exchange (ETDEWEB)

    Linke, Sara Y.S.; Tsiflikas, Ilias; Gatidis, Sergios; Schaefer, Juergen F. [University of Tuebingen, Department of Diagnostic and Interventional Radiology, Tuebingen (Germany); Herz, Klaus [University of Tuebingen, Department for Radiation Protection, Tuebingen (Germany); Szavay, Phillipp [Lucerne Cantonal Hospital, Pediatric Surgery, Lucerne 16 (Switzerland)

    2016-06-15

    To assess dose area products (DAP) and effective doses (ED) of voiding cystourethrography (VCUG) in children using optimized protocols on a modern flat detector unit. DAP and ED were evaluated in 651 VCUG (316 girls, median age: 2.25 years) between 2009 and 2012. DAP was analyzed in relation to patient characteristics (gender, age, presence of pathological findings) and experience of performing physician using analysis of variance. ED values were estimated using adapted conversion factors from the literature. Diagnostic image quality was validated by two experienced physicians using a 3-point scale. Median DAP/ED was 0.5 cGycm{sup 2}/4.56 μSv (boys: 0.6 cGycm{sup 2}/6.16 μSv; girls: 0.4 cGycm{sup 2}/3.54 μSv). In 300 studies without pathologic findings DAP was 0.35 cGycm{sup 2}, whereas 351 studies with pathologic findings had a median DAP of 0.7 cGycm{sup 2}. No significant relationship between DAP and experience of radiologist was observed. Image validation resulted in an overall good to excellent rating. DAP and ED can be markedly reduced in paediatric VCUG performed with optimized protocols on modern equipment without a noticeable decrease in diagnostic image quality. (orig.)

  2. A Study on Control System Design Based on ARM Sea Target Search System

    Directory of Open Access Journals (Sweden)

    Lin Xinwei

    2015-01-01

    Full Text Available The infrared detector is used for sea target search, which can assist humans in searching suspicious objects at night and under poor visibility conditions, and improving search efficiency. This paper applies for interrupt and stack technology to solve problems of data losses that may be caused by one-to-many multi-byte protocol communication. Meanwhile, this paper implements hardware and software design of the system based on industrial-grade ARM control chip and uC / OS-II embedded operating system. The control system in the sea target search system is an information exchange and control center of the whole system, which solves the problem of controlling over the shooting angle of the infrared detector in the process of target search. After testing, the control system operates stably and reliably, and realizes rotation and control functions of the pan/tilt platform during automatic search, manual search and track.

  3. Co-verification of hardware and software for ARM SoC design

    CERN Document Server

    Andrews, Jason

    2004-01-01

    Hardware/software co-verification is how to make sure that embedded system software works correctly with the hardware, and that the hardware has been properly designed to run the software successfully -before large sums are spent on prototypes or manufacturing. This is the first book to apply this verification technique to the rapidly growing field of embedded systems-on-a-chip(SoC). As traditional embedded system design evolves into single-chip design, embedded engineers must be armed with the necessary information to make educated decisions about which tools and methodology to deploy. SoC verification requires a mix of expertise from the disciplines of microprocessor and computer architecture, logic design and simulation, and C and Assembly language embedded software. Until now, the relevant information on how it all fits together has not been available. Andrews, a recognized expert, provides in-depth information about how co-verification really works, how to be successful using it, and pitfalls to avoid. H...

  4. An augmented reality C-arm for intraoperative assessment of the mechanical axis: a preclinical study.

    Science.gov (United States)

    Fallavollita, Pascal; Brand, Alexander; Wang, Lejing; Euler, Ekkehard; Thaller, Peter; Navab, Nassir; Weidert, Simon

    2016-11-01

    Determination of lower limb alignment is a prerequisite for successful orthopedic surgical treatment. Traditional methods include the electrocautery cord, alignment rod, or axis board which rely solely on C-arm fluoroscopy navigation and are radiation intensive. To assess a new augmented reality technology in determining lower limb alignment. A camera-augmented mobile C-arm (CamC) technology was used to create a panorama image consisting of hip, knee, and ankle X-rays. Twenty-five human cadaver legs were used for validation with random varus or valgus deformations. Five clinicians performed experiments that consisted in achieving acceptable mechanical axis deviation. The applicability of the CamC technology was assessed with direct comparison to ground-truth CT. A t test, Pearson's correlation, and ANOVA were used to determine statistical significance. The value of Pearson's correlation coefficient R was 0.979 which demonstrates a strong positive correlation between the CamC and ground-truth CT data. The analysis of variance produced a p value equal to 0.911 signifying that clinician expertise differences were not significant with regard to the type of system used to assess mechanical axis deviation. All described measurements demonstrated valid measurement of lower limb alignment. With minimal effort, clinicians required only 3 X-ray image acquisitions using the augmented reality technology to achieve reliable mechanical axis deviation.

  5. Recoil detector test for the day-one experiment at HESR

    Energy Technology Data Exchange (ETDEWEB)

    Hu, Qiang [Institute of Modern Physics, CAS, 730000 Lanzhou (China); Forschungszentrum Juelich, 52425 Juelich (Germany); Xu, Huagen; Ritman, James [Forschungszentrum Juelich, 52425 Juelich (Germany)

    2013-07-01

    The proposed day-one experiment at HESR is a dedicated measurement of antiproton-proton elastic scattering. The aim of the day-one experiment is to determine the elastic differential parameters (total cross section σ{sub T}, the ratio of real to imaginary part of the forward scattering amplitude ρ, and the slope parameter B) by measuring a large range of 4-momentum transfer squared t (0.0008-0.1 GeV{sup 2}). The conceptual design of the day-one experiment is to measure the elastic scattered antiproton and recoil proton, by a tracking detector in the small polar angle range and by an energy detector near 90 , respectively. The recoil arm covers a maximum polar angle range from 71 to 90 and consists of two silicon strip detectors (76.8(length) x 50.0(width) x 1.0(thickness) mm{sup 3}) and two germanium detectors (80.4(length) x 50.0(width) x 5.0 (11.0) (thickness) mm{sup 3}). All detectors are single sided structure with 1.2 mm pitch. The silicon detectors will be used to detect recoil protons with energy up to about 12 MeV and the germanium detectors will be used to detect protons with energy from 12 MeV to 60 MeV. At present, one recoil arm is being constructed and the test for the detectors with radioactive sources is on-going. Preliminary test results indicate that all detectors are operational and work properly. The latest test results of these detectors are presented.

  6. Usefulness of multi detector row computed tomography for detection of flat and depressed colorectal cancer

    International Nuclear Information System (INIS)

    Izumiya, Takashi; Hirata, Ichiro; Hamamoto, Norihiro; Matsuki, Mitsuru; Narabayashi, Isamu; Nishiguchi, Kanji; Okuda, Junji; Tanigawa, Nobuhiko; Katsu, Ken-ichi

    2005-01-01

    Recently, the clinical usefulness of colorectal cancer screening by CT colonography has been reported in Europe and the USA. However, in Japan, the diagnosis of flat or depressed colorectal cancer lesions has been emphasized, and the question of whether CT colonography facilitates visualization of these lesions remains to be answered. In the present study, we compared the visualization of flat and depressed colorectal cancer lesions by CT colonography with that of protruding lesions. We investigated 33 Dukes A colorectal cancer lesions that had been examined by 3D-CT, colonoscopy, and barium enema prior to surgery. In all patients, CT colonography was performed immediately after colonoscopy. Volume rendering was used for 3-D rearrangement, and imaging findings were examined with respect to morphology, tumor diameter, and tumor height. All (14/14) of the protruding-type lesions were visualized by CT colonography, whereas 78.9% (15/19) of the flat and depressed-type lesions were visualized. There was no significant difference in tumor diameter between protruding-type lesions and flat and depressed-type lesions. With respect to tumor height, 100% of the lesions measuring 2 mm or more in height were visualized, whereas only 42.9% of those measuring less than 2 mm in height were visualized; the difference was significant (P<0.001). These results suggest that the visualization capacity of CT colonography is associated with tumor height, but not with tumor diameter. Currently, lesions measuring 2 mm or more in height can be visualized reliably by CT colonography. (authors)

  7. Influence of arm positioning on tomographic thallium-201 myocardial perfusion imaging and the effect of attenuation correction

    International Nuclear Information System (INIS)

    Prvulovich, E.M.; Jarritt, P.H.; Vorontsova, E.; Bomanji, J.B.; Ell, P.J.

    2000-01-01

    Lateral attenuation in single-photon emission tomography (SPET) myocardial perfusion imaging (MPI) has been attributed to the left arm if it is held by the patient's side during data acquisition. As a result MPI data are conventionally acquired with the arms held above the head. The aims of this study were to determine the effect of imaging arms down on reconstructed tomographic images depicting regional myocardial thallium-201 distribution and to assess whether attenuation-corrected (AC) myocardial perfusion images acquired arms down could replace uncorrected (NC) images acquired arms up for routine clinical service. Twenty-eight patients referred for routine MPI underwent sequential 180 emission/transmission imaging for attenuation correction using an L-shaped dual-headed gamma camera (GE Optima) fitted with two gadolinium-153 scanning line sources. Delay data were acquired twice: once supine with the arms up and then supine with the arms down. Detector radius of rotation (ROR) for arms up and arms-down studies was recorded. For each data set, count density was measured in 17 segments of a polar plot and segmental uptake expressed relative to study maximum. Oblique images were assessed qualitatively by two observers blinded to study type for tracer distribution and overall quality. Transmission maps were assessed for truncation. Mean detector ROR was 190 mm for arms-up studies and 232 mm for arms-down studies (P 201 Tl distribution, particularly anterolaterally. There is lateral undercorrection in approximately 10% of AC arms-down studies, possibly because of attenuation map truncation. Image quality is reduced in about one-third of AC arms-down studies compared with NC arms-up studies. These data suggest that this attenuation correction method is not sufficiently robust to allow routine acquisition of MPI data with the arms down. (orig.)

  8. Optimization of detector pixel size for stent visualization in x-ray fluoroscopy

    International Nuclear Information System (INIS)

    Jiang Yuhao; Wilson, David L.

    2006-01-01

    Pixel size is of great interest in the flat-panel detector design because of its potential impact on image quality. In the particular case of angiographic x-ray fluoroscopy, small pixels are required in order to adequately visualize interventional devices such as guidewires and stents which have wire diameters as small as 200 and 50 μm, respectively. We used quantitative experimental and modeling techniques to investigate the optimal pixel size for imaging stents. Image quality was evaluated by the ability of subjects to perform two tasks: detect the presence of a stent and discriminate a partially deployed stent from a fully deployed one in synthetic images. With measurements at 50, 100, 200, and 300 μm, the 100 μm pixel size gave the maximum contrast sensitivity for the detection experiment with the idealized direct detector. For an idealized indirect detector with a scintillating layer, an optimal pixel size was obtained at 200 μm pixel size. A channelized human observer model predicted a peak at 150 and 170 μm, for the idealized direct and indirect detectors, respectively. With regard to the stent deployment task for both detector types, smaller pixel sizes are favored and there is a steep drop in performance with larger pixels. In general, with the increasing exposures, the model and measurements give the enhanced contrast sensitivities and a smaller optimal pixel size. The effects of electronic noise and fill factor were investigated using the model. We believe that the experimental results and human observer model predications can help guide the flat-panel detector design. In addition, the human observer model should work on the similar images and be applicable to the future model and actual flat-panel implementations

  9. Owhership of flats

    OpenAIRE

    Přibil, Jan

    2012-01-01

    Ownership of Flats Summary In his diploma thesis "Ownership of Flats", the author focuses on applicable law of flat ownership in the Czech Republic, especially the Flat Ownership Act 72/1994 Sb. The author puts the contemporary regulation in historical context; he describes the theoretical principles underlining the current applicable law and defines in detail the basic legal terms used by the Flat Ownership Act. Original and derivative forms of flat ownership acquisition are explained, namel...

  10. Permitting plan for Project W-340, Tank 241-C-106 manipulator retrieval arm

    International Nuclear Information System (INIS)

    Tollefson, K.S.

    1995-01-01

    This document describes the regulatory requirements and describes alternative strategies for obtaining permits and approvals for Project W-340, Tank 241-C-106 Manipulator Retrieval Arm. A comprehensive review of environmental regulations has indicated that several environmental reviews, permits, and approvals are required before design, construction, and operation of the facility. The environmental reviews, permits, and approvals, as well the regulatory authority potentially applicable to the Project W-340 Long Reach Manipulator Arm include the following: National Environmental Policy Act of 1969 -- US Department of Energy, Headquarters; State Environmental Policy Act of 1971 -- State of Washington Department of Ecology; Air Permitting; Dangerous Waste Permitting; Miscellaneous Reviews/Permits/Approvals. This document describes the environmental reviews, permits, and approval requirements for the project. It provides a summary of permit application data requirements, alternative strategies for permit completion and approval, as well as the estimated probability of success for each alternative strategy

  11. Comparative evaluations of surface contamination detectors calibration with radioactive sources - used in the Goiania accident, and standard sources

    International Nuclear Information System (INIS)

    Becker, P.H.B.; Marecha, M.H.H.

    1997-01-01

    The construction of Cs-137 standard flat sources for calibration of surface contamination detectors, used in the Goiania accident in 1987, is described and the procedures adopted are reported. At that time, standard sources were not available. Nowadays the Instituto de Radioprotecao e Dosimetria has standard sources acquired from Amersham which are used as calibration standards for surface contamination detectors. Comparative evaluations between the standard flat sources constructed for the accident and the calibrated ones are presented

  12. Application of C-arm CT-guided targeted puncturing technique in performing non-vascular interventional biopsy or interventional therapy

    International Nuclear Information System (INIS)

    Li Zhen; Han Xinwei; Jiao Dechao; Ren Jianzhuang; Su Yu; Ye Hui

    2011-01-01

    Objective: to investigate the clinical value of C-arm CT-guided targeted puncturing technique in performing non, vascular interventional biopsy or interventional therapy. Methods: Thirty, one patients, who were encountered in authors' hospital during the period from July 2010 to September 2010, were involved in this study. C-arm CT-guided percutaneous targeted puncturing biopsy or interventional therapy was performed in all 31 patients. All patients had complete clinical data. The complications and positive rate of biopsy were recorded and analyzed. Results: Under C-arm CT-guidance, percutaneous interventional therapy was carried out in 13 patients. The interventional procedures included radiofrequency ablation therapy for hepatic cellular carcinoma (n=2), pelvic abscess draining (n=1), hepatic abscess draining (n=1), ethanol injection for liver cancer (n=4), sclerotic therapy with ethanol injection for renal cyst (n=2), sclerotic therapy with ethanol injection for liver cyst (n=2) and catheter-indwelling drainage for pancreatic pseudocyst (n=1). percutaneous interventional biopsy was performed in the remaining 18 cases, including liver (n=4), lung (n=7), mediastinum (n=2), bone and soft tissue (n=4) and neck mass (n=1). All the procedures were successfully accomplished, no technique, related complications occurred during the operation. For biopsy examination in 18 cases, the positive rate was 94.4% (17/18) and false, negative results was seen in one case with lung lesion. Conclusion: The percutaneous targeted puncturing technique with C, arm CT-guidance combines the advantages of both CT scanning and fluoroscopy. The use of real, time road, mapping function can effectively guide the puncturing and therapeutic management, which can not only optimize the workflow, save the operation time, but also improve the success rate and technical safety. Therefore, it is of great value to popularize this targeted puncturing technique. (authors)

  13. Radiation protection for an intraoperative X-ray source compared to C-arm fluoroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Schneider, Frank; Clausen, Sven; Jahnke, Anika; Steil, Volker; Wenz, Frederik [Heidelberg Univ., University Medical Center Mannheim (Germany). Dept. of Radiation Oncology; Bludau, Frederic; Obertacke, Udo [Heidelberg Univ., University Medical Center Mannheim (Germany). Dept. of Trauma Surgery; Suetterlin, Marc [Heidelberg Univ., University Medical Center Mannheim (Germany). Dept. of Obstetrics and Gynaecology

    2014-10-01

    Background: Intraoperative radiotherapy (IORT) using the INTRABEAM {sup registered} system promises a flexible use regarding radiation protection compared to other approaches such as electron treatment or HDR brachytherapy with {sup 192}Ir or {sup 60}Co. In this study we compared dose rate measurements of breast- and Kypho-IORT with C-arm fluoroscopy which is needed to estimate radiation protection areas. Materials and Methods: C-arm fluoroscopy, breast- and Kypho-IORTs were performed using phantoms (silicon breast or bucket of water). Dose rates were measured at the phantom's surface, at 30 cm, 100 cm and 200 cm distance. Those measurements were confirmed during 10 Kypho-IORT and 10 breast-IORT patient treatments. Results: The measured dose rates were in the same magnitude for all three paradigms and ranges from 20 μSv/h during a simulated breast-IORT at two meter distance up to 64 mSv/h directly at the surface of a simulated Kypho-IORT. Those measurements result in a circle of controlled area (yearly doses > 6 mSv) for each paradigm of about 4 m ± 2 m. Discussion/Conclusions: All three paradigms show comparable dose rates which implies that the radiation protection is straight forward and confirms the flexible use of the INTRABEAM {sup registered} system. (orig.)

  14. C-arm Cone Beam Computed Tomography: A New Tool in the Interventional Suite.

    Science.gov (United States)

    Raj, Santhosh; Irani, Farah Gillan; Tay, Kiang Hiong; Tan, Bien Soo

    2013-11-01

    C-arm Cone Beam CT (CBCT) is a technology that is being integrated into many of the newer angiography systems in the interventional suite. Due to its ability to provide cross sectional imaging, it has opened a myriad of opportunities for creating new clinical applications. We review the technical aspects, current reported clinical applications and potential benefits of this technology. Searches were made via PubMed using the string "CBCT", "Cone Beam CT", "Cone Beam Computed Tomography" and "C-arm Cone Beam Computed Tomography". All relevant articles in the results were reviewed. CBCT clinical applications have been reported in both vascular and non-vascular interventions. They encompass many aspects of a procedure including preprocedural planning, intraprocedural guidance and postprocedural assessment. As a result, they have allowed the interventionalist to be safer and more accurate in performing image guided procedures. There are however several technical limitations. The quality of images produced is not comparable to conventional computed tomography (CT). Radiation doses are also difficult to quantify when compared to CT and fluoroscopy. CBCT technology in the interventional suite has contributed significant benefits to the patient despite its current limitations. It is a tool that will evolve and potentially become an integral part of imaging guidance for intervention.

  15. Effects induced by high and low intensity laser plasma on SiC Schottky detectors

    Directory of Open Access Journals (Sweden)

    Sciuto Antonella

    2018-01-01

    Full Text Available Silicon-Carbide detectors are extensively employed as diagnostic devices in laser-generated plasma, allowing the simultaneous detection of photons, electrons and ions, when used in time-of-flight configuration. The plasma generated by high intensity laser (1016 W/cm2 producing high energy ions was characterized by SiC detector with a continuous front-electrode, and a very thick active depth, while SiC detector with an Interdigit front-electrode was used to measure the low energy ions of plasma generated by low intensity laser (1010 W/cm2. Information about ion energy, number of charge states, plasma temperature can be accurately obtained. However, laser exposure induces the formation of surface and bulk defects whose concentration increases with increasing the time to plasma exposure. The surface defects consist of clusters with a main size of the order of some microns and they modify the diode barrier height and the efficiency of the detector as checked by alpha spectrometry. The bulk defects, due to the energy loss of detected ions, strongly affect the electrical properties of the device, inducing a relevant increase of the leakage (reverse current and decrease the forward current related to a deactivation of the dopant in the active detector region.

  16. Effects induced by high and low intensity laser plasma on SiC Schottky detectors

    Science.gov (United States)

    Sciuto, Antonella; Torrisi, Lorenzo; Cannavò, Antonino; Mazzillo, Massimo; Calcagno, Lucia

    2018-01-01

    Silicon-Carbide detectors are extensively employed as diagnostic devices in laser-generated plasma, allowing the simultaneous detection of photons, electrons and ions, when used in time-of-flight configuration. The plasma generated by high intensity laser (1016 W/cm2) producing high energy ions was characterized by SiC detector with a continuous front-electrode, and a very thick active depth, while SiC detector with an Interdigit front-electrode was used to measure the low energy ions of plasma generated by low intensity laser (1010 W/cm2). Information about ion energy, number of charge states, plasma temperature can be accurately obtained. However, laser exposure induces the formation of surface and bulk defects whose concentration increases with increasing the time to plasma exposure. The surface defects consist of clusters with a main size of the order of some microns and they modify the diode barrier height and the efficiency of the detector as checked by alpha spectrometry. The bulk defects, due to the energy loss of detected ions, strongly affect the electrical properties of the device, inducing a relevant increase of the leakage (reverse) current and decrease the forward current related to a deactivation of the dopant in the active detector region.

  17. Arm-length stabilisation for interferometric gravitational-wave detectors using frequency-doubled auxiliary lasers

    OpenAIRE

    Mullavey, Adam J.; Slagmolen, Bram J. J.; Miller, John; Evans, Matthew; Fritschel, Peter; Sigg, Daniel; Waldman, Sam J.; Shaddock, Daniel A.; McClelland, David E.

    2011-01-01

    Residual motion of the arm cavity mirrors is expected to prove one of the principal impediments to systematic lock acquisition in advanced gravitational-wave interferometers. We present a technique which overcomes this problem by employing auxiliary lasers at twice the fundamental measurement frequency to pre-stabilise the arm cavities’ lengths. Applying this approach, we reduce the apparent length noise of a 1.3 m long, independently suspended Fabry-Perot cavity to 30 pm rms and successfully...

  18. Sensitization by UV light of α-Al2O3:C polycrystalline detectors

    International Nuclear Information System (INIS)

    Meira B, L. C.; Rubio F, H.; Neres de A, E.; Santos, A.

    2014-08-01

    This paper describes an increase in sensitivity to gamma and beta radiation on α-Al 2 O 3 :C polycrystalline detector, which has been produced by a sol-gel process, following previous exposure to ultraviolet light. The increased sensitivity of the detector as a function of the exposure time and ultraviolet wavelength was studied. Since the main luminescent centers have emission peaks at different wavelengths, selective measurements of thermoluminescent emission intensity were done, in order to investigate the possible conversion of centers as a result of the exposition to ultraviolet light. Experimental results indicate that the nature and parameters of the luminescent centers in α-Al 2 O 3 :C sol-gel material can be very different of those in α-Al 2 O 3 :C single crystal. (author)

  19. Spine segmentation from C-arm CT data sets: application to region-of-interest volumes for spinal interventions

    Science.gov (United States)

    Buerger, C.; Lorenz, C.; Babic, D.; Hoppenbrouwers, J.; Homan, R.; Nachabe, R.; Racadio, J. M.; Grass, M.

    2017-03-01

    Spinal fusion is a common procedure to stabilize the spinal column by fixating parts of the spine. In such procedures, metal screws are inserted through the patients back into a vertebra, and the screws of adjacent vertebrae are connected by metal rods to generate a fixed bridge. In these procedures, 3D image guidance for intervention planning and outcome control is required. Here, for anatomical guidance, an automated approach for vertebra segmentation from C-arm CT images of the spine is introduced and evaluated. As a prerequisite, 3D C-arm CT images are acquired covering the vertebrae of interest. An automatic model-based segmentation approach is applied to delineate the outline of the vertebrae of interest. The segmentation approach is based on 24 partial models of the cervical, thoracic and lumbar vertebrae which aggregate information about (i) the basic shape itself, (ii) trained features for image based adaptation, and (iii) potential shape variations. Since the volume data sets generated by the C-arm system are limited to a certain region of the spine the target vertebra and hence initial model position is assigned interactively. The approach was trained and tested on 21 human cadaver scans. A 3-fold cross validation to ground truth annotations yields overall mean segmentation errors of 0.5 mm for T1 to 1.1 mm for C6. The results are promising and show potential to support the clinician in pedicle screw path and rod planning to allow accurate and reproducible insertions.

  20. Modeling and simulation of critical parameters of the first chamber of the dimuon arm spectrometer of the Alice experiment

    International Nuclear Information System (INIS)

    Guez, D.

    2003-10-01

    The Alice experiment that is dedicated to the study of ultra-relativistic heavy ion collisions, will take place in the future large hadron collider (LHC) at CERN. The dimuon arm spectrometer of the Alice experiment is devoted to the search of a new signature of the existence of the quark gluon plasma (QGP). The first chapter is dedicated to the physics notions linked to the study of QGP, a few signatures are proposed for the detection of QGP, particularly the signature concerning the production rate of quarkonium. The second chapter deals with particle detection involved in Alice experiment, the dimuon arm spectrometer is a detector dedicated to the track reconstruction of muons issued from the decay of heavy mesons from J/Ψ and Υ families. The third and the fourth chapters present the studies made to integrate a reliable model of the dimuon arm in the global simulation code of Alice (Aliroot). The fifth chapter presents the software TB 2 that has been developed within the framework of this thesis in order to check and control the output data when the detector is tested with a real particle beam. The sixth chapter presents the results of the tests that have been performed with a 7 GeV/c pion beam. These tests have shown that the electronic noise is coherent with the specifications of Alice experiment. A factor 1,8 between the highest and the weakest values of the gain has been measured in the chamber. The detection efficiency of the chamber has been estimated to 99% in the different cases studied. (A.C.)

  1. Simultaneous 3D–2D image registration and C-arm calibration: Application to endovascular image-guided interventions

    Energy Technology Data Exchange (ETDEWEB)

    Mitrović, Uroš [Faculty of Electrical Engineering, University of Ljubljana, Tržaška 25, Ljubljana 1000, Slovenia and Cosylab, Control System Laboratory, Teslova ulica 30, Ljubljana 1000 (Slovenia); Pernuš, Franjo [Faculty of Electrical Engineering, University of Ljubljana, Tržaška 25, Ljubljana 1000 (Slovenia); Likar, Boštjan; Špiclin, Žiga, E-mail: ziga.spiclin@fe.uni-lj.si [Faculty of Electrical Engineering, University of Ljubljana, Tržaška 25, Ljubljana 1000, Slovenia and Sensum, Computer Vision Systems, Tehnološki Park 21, Ljubljana 1000 (Slovenia)

    2015-11-15

    Purpose: Three-dimensional to two-dimensional (3D–2D) image registration is a key to fusion and simultaneous visualization of valuable information contained in 3D pre-interventional and 2D intra-interventional images with the final goal of image guidance of a procedure. In this paper, the authors focus on 3D–2D image registration within the context of intracranial endovascular image-guided interventions (EIGIs), where the 3D and 2D images are generally acquired with the same C-arm system. The accuracy and robustness of any 3D–2D registration method, to be used in a clinical setting, is influenced by (1) the method itself, (2) uncertainty of initial pose of the 3D image from which registration starts, (3) uncertainty of C-arm’s geometry and pose, and (4) the number of 2D intra-interventional images used for registration, which is generally one and at most two. The study of these influences requires rigorous and objective validation of any 3D–2D registration method against a highly accurate reference or “gold standard” registration, performed on clinical image datasets acquired in the context of the intervention. Methods: The registration process is split into two sequential, i.e., initial and final, registration stages. The initial stage is either machine-based or template matching. The latter aims to reduce possibly large in-plane translation errors by matching a projection of the 3D vessel model and 2D image. In the final registration stage, four state-of-the-art intrinsic image-based 3D–2D registration methods, which involve simultaneous refinement of rigid-body and C-arm parameters, are evaluated. For objective validation, the authors acquired an image database of 15 patients undergoing cerebral EIGI, for which accurate gold standard registrations were established by fiducial marker coregistration. Results: Based on target registration error, the obtained success rates of 3D to a single 2D image registration after initial machine-based and

  2. Equipment for electrochemical etching of dielectric track detectors

    International Nuclear Information System (INIS)

    Turek, K.; Novak, M.

    1992-01-01

    The facility is designed for electrochemical etching of solid state track detectors, devised for thicknesses in excess of 200 μm as employed for direct detection of charged particles or neutrons. The device consists of a high-voltage a.c. supply for the electrodes and an assembly whose body consists of a flat tank, on the surface of which is formed a channel for feeding the heating and cooling media. The tank is covered by a gasket, an earthed metal plate for the detector, a mask with holes determining the shape of the etched area, and a pressure plate. The pressure plate is fitted with a system of holes for the etching solution. Needle-shaped high-voltage electrodes are accommodated in the holes of the mask. The underlying principle of the invention consists in the fact that a rubber pad with guide holes for the needle-shaped electrodes lies on the pressure plate. Each electrode is composed of a central tip and an outer jacket, separated by an insulating layer; connection is provided by a light emitting diode. (Z.S.). 2 figs

  3. Neutron measurements on the JET tokamak by means of bubble detectors

    International Nuclear Information System (INIS)

    Gherendi, M.; Craciunescu, T.; Pantea, A.; Zoita, V.; Edlington, T.; Kiptily, V.; Popovichev, S.; Murari, A.

    2009-01-01

    Full text: The bubble detectors (superheated fluid detectors - SHFDs) are based on suspensions of superheated fluid droplets which vaporise into bubbles when nucleated by radiation interactions. The active detecting medium is in the form of microscopic (20-50 μm) droplets suspended within an elastic polymer. The bubble detectors are of interest for neutron detection in nuclear fusion devices due to some particular characteristics: - High neutron detection efficiency (counts/unit fluence) that ranges from about 4x10 -2 to 4x10 -5 ; - Almost flat, threshold-type energy response over a broad energy range (10's keV to 10's MeV); - The possibility of having any energy threshold within the above-mentioned energy range; - Practically zero sensitivity to gamma-radiation; - Good spatial resolution (sub-centimetre resolution in the image plane). A series of the neutron measurements have been carried out by means of bubble detectors on the JET tokamak, at Culham Science Centre, Abingdon, UK, during the experimental campaigns C17-C26 (2007-2009). The neutron field parameters (yield, fluence, energy distribution) at a specific location outside the JET Torus Hall have been measured using three types of bubble detectors (BD-PND, DEFENDER, and BDS). The bubble detector measurement location is situated at the end of a vertical collimated line of sight, behind the TOFOR neutron time-of-flight spectrometer. The field-of-view is defined by a variable pre-collimator located on top of the JET tokamak. This paper reports only on the neutron fluence measurements. Spatial (radial and toroidal) distributions of the neutron fluence have been obtained with a two-dimensional array having up to 30 bubble detectors. The operation of the bubble detector array as a neutron pinhole camera having a radial resolution at the JET vacuum chamber mid-plane of about 55 mm was demonstrated in measurements using various openings of the pre-collimator. (authors)

  4. The MVACS Robotic Arm Camera

    Science.gov (United States)

    Keller, H. U.; Hartwig, H.; Kramm, R.; Koschny, D.; Markiewicz, W. J.; Thomas, N.; Fernades, M.; Smith, P. H.; Reynolds, R.; Lemmon, M. T.; Weinberg, J.; Marcialis, R.; Tanner, R.; Boss, B. J.; Oquest, C.; Paige, D. A.

    2001-08-01

    The Robotic Arm Camera (RAC) is one of the key instruments newly developed for the Mars Volatiles and Climate Surveyor payload of the Mars Polar Lander. This lightweight instrument employs a front lens with variable focus range and takes images at distances from 11 mm (image scale 1:1) to infinity. Color images with a resolution of better than 50 μm can be obtained to characterize the Martian soil. Spectral information of nearby objects is retrieved through illumination with blue, green, and red lamp sets. The design and performance of the camera are described in relation to the science objectives and operation. The RAC uses the same CCD detector array as the Surface Stereo Imager and shares the readout electronics with this camera. The RAC is mounted at the wrist of the Robotic Arm and can characterize the contents of the scoop, the samples of soil fed to the Thermal Evolved Gas Analyzer, the Martian surface in the vicinity of the lander, and the interior of trenches dug out by the Robotic Arm. It can also be used to take panoramic images and to retrieve stereo information with an effective baseline surpassing that of the Surface Stereo Imager by about a factor of 3.

  5. High-Resolution Fast-Neutron Spectrometry for Arms Control and Treaty Verification

    Energy Technology Data Exchange (ETDEWEB)

    David L. Chichester; James T. Johnson; Edward H. Seabury

    2012-07-01

    Many nondestructive nuclear analysis techniques have been developed to support the measurement needs of arms control and treaty verification, including gross photon and neutron counting, low- and high-resolution gamma spectrometry, time-correlated neutron measurements, and photon and neutron imaging. One notable measurement technique that has not been extensively studied to date for these applications is high-resolution fast-neutron spectrometry (HRFNS). Applied for arms control and treaty verification, HRFNS has the potential to serve as a complimentary measurement approach to these other techniques by providing a means to either qualitatively or quantitatively determine the composition and thickness of non-nuclear materials surrounding neutron-emitting materials. The technique uses the normally-occurring neutrons present in arms control and treaty verification objects of interest as an internal source of neutrons for performing active-interrogation transmission measurements. Most low-Z nuclei of interest for arms control and treaty verification, including 9Be, 12C, 14N, and 16O, possess fast-neutron resonance features in their absorption cross sections in the 0.5- to 5-MeV energy range. Measuring the selective removal of source neutrons over this energy range, assuming for example a fission-spectrum starting distribution, may be used to estimate the stoichiometric composition of intervening materials between the neutron source and detector. At a simpler level, determination of the emitted fast-neutron spectrum may be used for fingerprinting 'known' assemblies for later use in template-matching tests. As with photon spectrometry, automated analysis of fast-neutron spectra may be performed to support decision making and reporting systems protected behind information barriers. This paper will report recent work at Idaho National Laboratory to explore the feasibility of using HRFNS for arms control and treaty verification applications, including simulations

  6. ''VECTON-1'' two-arm spectrometer for rho--meson photoproduction study

    International Nuclear Information System (INIS)

    Anokhin, M.V.; Kanetsyan, A.R.; Kukarev, V.M.

    1977-01-01

    A two-arm spectrometer for registering p - mesons according to a charged pion and one of the neutral pion disintegration photons has been designed. The spectrometer arm which registers charged pion comprises a deflecting magnet, dual wide-gap spark chambers and dual scintillation counters. The spectrometer arm for registering the disintegration photon comprises a scintillation counter, a dual wide-gap spark chamber and a shower detector. The principal characteristics of the spectrometer components are listed. The functional diagram of the useful events selection unit is discussed. It is shown that the experimental results obtained with the aid of the ''VEKTON-1'' set-up are in good agreement with the earlier results: P - meson rest energy is 731+-30 MeV, the resonant range width is 195+-58 MeV

  7. A new system for fully automatic inspection of digital flat-panel detector radiographs of aluminium castings

    International Nuclear Information System (INIS)

    Fuchs, T.; Hassler, U.; Huetten, U.; Wenzel, T.

    2006-01-01

    The aim of our work was the integration of various newly-developed methods into a system for fully automatic radioscopic inspection of arbitrary casting parts. Using a 16-bit flat-panel detector, projections in arbitrary directions through the part are acquired and analysed. The software tool for inspection can be separated into five stages: registration, calibration, image processing, fault segmentation, and quality assessment. Thereby, each step is realized with full 16-bit data processing. Within the first processing stage, information about the physical length and density of the aluminium structures is extracted from the primary projections. Next, the primary image is registered with a reference image, which was acquired previously. Afterwards, the third stage combines both reference image-based and reference-less testing. A filter is applied, which adapts automatically to the local object structure by referring to the properties of the reference. Thereby, the self-adapting filter selects its size, direction and filter method optimally according to the local situation. Similar to the reference-less procedure, a subtraction is followed by a threshold operation, resulting in a map of regions that are suspected to be faulty. The fourth step aims at an elimination of false-positive detections. Again, two methods are applied successively: evaluation of local image features at suspicious positions and a classification based on teachings independent of position and orientation of the faults. Within the last step the quality criteria are applied. These criteria may concern fault size and depth, the density of faults in critical regions and a minimum distance between two or more faults. (orig.)

  8. Development of a simple detector response function generation program: The CEARDRFs code

    International Nuclear Information System (INIS)

    Wang Jiaxin; Wang Zhijian; Peeples, Johanna; Yu Huawei; Gardner, Robin P.

    2012-01-01

    A simple Monte Carlo program named CEARDRFs has been developed to generate very accurate detector response functions (DRFs) for scintillation detectors. It utilizes relatively rigorous gamma-ray transport with simple electron transport, and accounts for two phenomena that have rarely been treated: scintillator non-linearity and the variable flat continuum part of the DRF. It has been proven that these physics and treatments work well for 3×3″ and 6×6″ cylindrical NaI detector in CEAR's previous work. Now this approach has been expanded to cover more scintillation detectors with various common shapes and sizes. Benchmark experiments of 2×2″ cylindrical BGO detector and 2×4×16″ rectangular NaI detector have been carried out at CEAR with various radiactive sources. The simulation results of CEARDRFs have also been compared with MCNP5 calculations. The benchmark and comparison show that CEARDRFs can generate very accurate DRFs (more accurate than MCNP5) at a very fast speed (hundred times faster than MCNP5). The use of this program can significantly increase the accuracy of applications relying on detector spectroscopy like prompt gamma-ray neutron activation analysis, X-ray fluorescence analysis, oil well logging and homeland security. - Highlights: ► CEARDRF has been developed to generate detector response functions (DRFs) for scintillation detectors a. ► Generated DRFs are very accurate. ► Simulation speed is hundreds of times faster than MCNP5. ► It utilizes rigorous gamma-ray transport with simple electron transport. ► It also accounts for scintillator non-linearity and the variable flat continuum part.

  9. Rietveld analysis using powder diffraction data with anomalous scattering effect obtained by focused beam flat sample method

    International Nuclear Information System (INIS)

    Tanaka, Masahiko; Katsuya, Yoshio; Sakata, Osami

    2016-01-01

    Focused-beam flat-sample method (FFM) is a new trial for synchrotron powder diffraction method, which is a combination of beam focusing optics, flat shape powder sample and area detectors. The method has advantages for X-ray diffraction experiments applying anomalous scattering effect (anomalous diffraction), because of 1. Absorption correction without approximation, 2. High intensity X-rays of focused incident beams and high signal noise ratio of diffracted X-rays 3. Rapid data collection with area detectors. We applied the FFM to anomalous diffraction experiments and collected synchrotron X-ray powder diffraction data of CoFe_2O_4 (inverse spinel structure) using X-rays near Fe K absorption edge, which can distinguish Co and Fe by anomalous scattering effect. We conducted Rietveld analyses with the obtained powder diffraction data and successfully determined the distribution of Co and Fe ions in CoFe_2O_4 crystal structure.

  10. Nuclear radiation detector based on ion implanted p-n junction in 4H-SiC

    International Nuclear Information System (INIS)

    Vervisch, V.; Issa, F.; Ottaviani, L.; Lazar, M.; Kuznetsov, A.; Szalkai, D.; Klix, A.; Lyoussi, A.; Vermeeren, L.; Hallen, A.

    2013-06-01

    In this paper, we propose a new device detector based on ion implanted p-n junction in 4H-SiC for nuclear instrumentation. We showed the interest to use 10 Boron as a Neutron Converter Layer in order to detect thermal neutrons. We present the main results obtained during irradiation tests performed in the Belgian Reactor 1. We show the capability of our detector by means of first results of the detector response at different reverse voltage biases and at different reactor power (authors)

  11. Detection of 14 MeV neutrons in high temperature environment up to 500 deg. C using 4H-SiC based diode detector

    Energy Technology Data Exchange (ETDEWEB)

    Szalkai, D.; Klix, A. [KIT- Karlsruhe Institute of Technology, Institute of Neutron Physics and Reactor Technology Karlsruhe 76344 (Germany); Ferone, R.; Issa, F.; Ottaviani, L.; Vervisch, V. [IM2NP, UMR CNRS 7334, Aix-Marseille University, Case 231 -13397 Marseille Cedex 20 (France); Gehre, D. [Inst. for Nucl.- and Particle-Phys., Dresden University of Technology, Dresden 01069 (Germany); Lyoussi, A. [CEA, DEN, Departement d' Etudes des Reacteurs, Service de Physique Experimentale, Laboratoire Dosimetrie Capteurs Instrumentation, 13108 Saint-Paul-lez-Durance (France)

    2015-07-01

    In reactor technology and industrial applications detection of fast and thermal neutrons plays a crucial role in getting relevant information about the reactor environment and neutron yield. The inevitable elevated temperatures make neutron yield measurements problematic. Out of the currently available semiconductors 4H-SiC seems to be the most suitable neutron detector material under extreme conditions due to its high heat and radiation resistance, large band-gap and lower cost of production than in case of competing diamond detectors. In the framework of the European I-Smart project, optimal {sup 4}H-SiC diode geometries were developed for high temperature neutron detection and have been tested with 14 MeV fast neutrons supplied by a deuterium-tritium neutron generator with an average neutron flux of 10{sup 10}-10{sup 11} n/(s*cm{sup 2}) at Neutron Laboratory of the Technical University of Dresden in Germany from room temperatures up to several hundred degrees Celsius. Based on the results of the diode measurements, detector geometries appear to play a crucial role for high temperature measurements up to 500 deg. C. Experimental set-ups using SiC detectors were constructed to simulate operation in the harsh environmental conditions found in the tritium breeding blanket of the ITER fusion reactor, which is planned to be the location of neutron flux characterization measurements in the near future. (authors)

  12. Optimization of dual-energy subtraction chest radiography by use of a direct-conversion flat-panel detector system.

    Science.gov (United States)

    Fukao, Mari; Kawamoto, Kiyosumi; Matsuzawa, Hiroaki; Honda, Osamu; Iwaki, Takeshi; Doi, Tsukasa

    2015-01-01

    We aimed to optimize the exposure conditions in the acquisition of soft-tissue images using dual-energy subtraction chest radiography with a direct-conversion flat-panel detector system. Two separate chest images were acquired at high- and low-energy exposures with standard or thick chest phantoms. The high-energy exposure was fixed at 120 kVp with the use of an auto-exposure control technique. For the low-energy exposure, the tube voltages and entrance surface doses ranged 40-80 kVp and 20-100 % of the dose required for high-energy exposure, respectively. Further, a repetitive processing algorithm was used for reduction of the image noise generated by the subtraction process. Seven radiology technicians ranked soft-tissue images, and these results were analyzed using the normalized-rank method. Images acquired at 60 kVp were of acceptable quality regardless of the entrance surface dose and phantom size. Using a repetitive processing algorithm, the minimum acceptable doses were reduced from 75 to 40 % for the standard phantom and to 50 % for the thick phantom. We determined that the optimum low-energy exposure was 60 kVp at 50 % of the dose required for the high-energy exposure. This allowed the simultaneous acquisition of standard radiographs and soft-tissue images at 1.5 times the dose required for a standard radiograph, which is significantly lower than the values reported previously.

  13. Non-destructive, preclinical evaluation of root canal anatomy of human teeth with flat-panel detector volume CT (FD-VCT)

    International Nuclear Information System (INIS)

    Heidrich, G.; Hassepass, F.; Dullin, C.; Grabbe, E.; Attin, T.; Hannig, C.

    2005-01-01

    Purpose: Successful endodontic diagnostics and therapy call for adequate depiction of the root canal anatomy with multimodal diagnostic imaging. The aim of the present study is to evaluate visualization of the endodont with flat-panel detector volume CT (FD-VCT). Materials and methods: 13 human teeth were examined with the prototype of a FD-VCT. After data acquisition and generation of volume data sets in volume rendering technology (VRT), the findings obtained were compared to conventional X-rays and cross-section preparations of the teeth. Results: The anatomical structures of the endodont such as root canals, side canals and communications between different root canals as well as dentricles could be detected precisely with FD-VCT. The length of curved root canals was also determined accurately. The spatial resolution of the system is around 140 μm. Only around 73% of the main root canals detected with FD-VCT and 87% of the roots could be visualized with conventional dental X-rays. None of the side canals, shown with FD-VCT, was detectable on conventional X-rays. In all cases the enamel and dentin of the teeth could be well delineated. No differences in image quality could be discerned between stored and freshly extracted teeth, or between primary and adult teeth. (orig.)

  14. Method of estimating patient skin dose from dose displayed on medical X-ray equipment with flat panel detector

    International Nuclear Information System (INIS)

    Fukuda, Atsushi; Koshida, Kichiro; Togashi, Atsuhiko; Matsubara, Kousuke

    2004-01-01

    The International Electrotechnical Commission (IEC) has stipulated that medical X-ray equipment for interventional procedures must display radiation doses such as air kerma in free air at the interventional reference point and dose area product to establish radiation safety for patients (IEC 60601-2-43). However, it is necessary to estimate entrance skin dose for the patient from air kerma for an accurate risk assessment of radiation skin injury. To estimate entrance skin dose from displayed air kerma in free air at the interventional reference point, it is necessary to consider effective energy, the ratio of the mass-energy absorption coefficient for skin and air, and the backscatter factor. In addition, since automatic exposure control is installed in medical X-ray equipment with flat panel detectors, it is necessary to know the characteristics of control to estimate exposure dose. In order to calculate entrance skin dose under various conditions, we investigated clinical parameters such as tube voltage, tube current, pulse width, additional filter, and focal spot size, as functions of patient body size. We also measured the effective energy of X-ray exposure for the patient as a function of clinical parameter settings. We found that the conversion factor from air kerma in free air to entrance skin dose is about 1.4 for protection. (author)

  15. Effect of gold photocathode contamination on a flat spectral response X-ray diode

    Science.gov (United States)

    Wang, Kun-lun; Zhang, Si-qun; Zhou, Shao-tong; Huang, Xian-bin; Ren, Xiao-dong; Dan, Jia-kun; Xu, Qiang

    2018-03-01

    A detector with an approximately flat spectral response is important for diagnosing intense thermal X-ray flux. A flat-spectral-response X-ray diode (FSR-XRD) utilizes a gold photocathode X-ray diode and a specially configured gold filter to give rise to a nearly flat spectral response in the photon energy range of 100-4000 eV. It has been observed that the spectral responses of several FSR-XRDs changed after a few shots of z-pinch experiments on the Primary Test Stand facility. This paper presents an analysis of the changes by fitting the spectral responses of the gold photocathodes using a model with a free parameter which characterizes the thickness of the contamination. The spectral responses of FSR-XRDs were calibrated with synchrotron radiation, and several cleaning methods were tested with the calibration. Considering the results of model and cleaning, it may be anticipated that contamination was the major reason of the response changing. Contamination worsened the flatness of the spectral response of the FSR-XRD and decreased the averaged response, hence it is important to avoid contamination. Current results indicate a requirement of further study of the contamination.

  16. Comparison of bulk and epitaxial 4H-SiC detectors for radiation hard particle tracking

    CERN Document Server

    Quinn, T; Bruzzi, M; Cunningham, W; Mathieson, K; Moll, M; Nelson, T; Nilsson, H E; Pintillie, I; Rahman, M; Reynolds, L; Sciortino, S; Sellin, P J; Strachan, H; Svensson, B G; Vaitkus, J

    2003-01-01

    Measurements and simulations have been carried out using bulk and epitaxial SiC detectors. Samples were irradiated to fluences of around 10**1**4 hardrons/cm**2. Material of thickness 40um gave a charge collection efficiency of 100% dropping to around 60% at 100mum thickness. Detailed MEDICI simulations incorporated the main defect levels in SiC, the vanadium center, Z-center and a mid-gap level as measured by deep level transient spectroscopy and other techniques. Calculated recombination currents and charge collection efficiencies at varying fluences were comparable to experimental data. The study suggests that SiC detectors will operate up to fluences around 10 **1**6/cm**2 as required by future particle physics experiments.

  17. Extended ellipse-line-ellipse trajectory for long-object cone-beam imaging with a mounted C-arm system

    International Nuclear Information System (INIS)

    Yu, Zhicong; Noo, Frédéric; Lauritsch, Günter; Dennerlein, Frank; Mao, Yanfei; Hornegger, Joachim

    2016-01-01

    Recent reports show that three-dimensional cone-beam (CB) imaging with a floor-mounted (or ceiling-mounted) C-arm system has become a valuable tool in interventional radiology. Currently, a circular short scan is used for data acquisition, which inevitably yields CB artifacts and a short coverage in the direction of the patient table. To overcome these two limitations, a more sophisticated data acquisition geometry is needed. This geometry should be complete in terms of Tuy’s condition and should allow continuous scanning, while being compatible with the mechanical constraints of mounted C-arm systems. Additionally, the geometry should allow accurate image reconstruction from truncated data. One way to ensure such a feature is to adopt a trajectory that provides full R-line coverage within the field-of-view (FOV). An R-line is any segment of line that connects two points on a source trajectory, and the R-line coverage is the set of points that belong to an R-line. In this work, we propose a novel geometry called the extended ellipse-line-ellipse (ELE) for long-object imaging with a mounted C-arm system. This trajectory is built from modules consisting of two elliptical arcs connected by a line. We demonstrate that the extended ELE can be configured in many ways so that full R-line coverage is guaranteed. Both tight and relaxed parametric settings are presented. All results are supported by extensive mathematical proofs provided in appendices. Our findings make the extended ELE trajectory attractive for axially-extended FOV imaging in interventional radiology. (paper)

  18. Psychophysical evaluation of the image quality of a dynamic flat-panel digital x-ray image detector using the threshold contrast detail detectability (TCDD) technique

    Science.gov (United States)

    Davies, Andrew G.; Cowen, Arnold R.; Bruijns, Tom J. C.

    1999-05-01

    We are currently in an era of active development of the digital X-ray imaging detectors that will serve the radiological communities in the new millennium. The rigorous comparative physical evaluations of such devices are therefore becoming increasingly important from both the technical and clinical perspectives. The authors have been actively involved in the evaluation of a clinical demonstration version of a flat-panel dynamic digital X-ray image detector (or FDXD). Results of objective physical evaluation of this device have been presented elsewhere at this conference. The imaging performance of FDXD under radiographic exposure conditions have been previously reported, and in this paper a psychophysical evaluation of the FDXD detector operating under continuous fluoroscopic conditions is presented. The evaluation technique employed was the threshold contrast detail detectability (TCDD) technique, which enables image quality to be measured on devices operating in the clinical environment. This approach addresses image quality in the context of both the image acquisition and display processes, and uses human observers to measure performance. The Leeds test objects TO[10] and TO[10+] were used to obtain comparative measurements of performance on the FDXD and two digital spot fluorography (DSF) systems, one utilizing a Plumbicon camera and the other a state of the art CCD camera. Measurements were taken at a range of detector entrance exposure rates, namely 6, 12, 25 and 50 (mu) R/s. In order to facilitate comparisons between the systems, all fluoroscopic image processing such as noise reduction algorithms, were disabled during the experiments. At the highest dose rate FDXD significantly outperformed the DSF comparison systems in the TCDD comparisons. At 25 and 12 (mu) R/s all three-systems performed in an equivalent manner and at the lowest exposure rate FDXD was inferior to the two DSF systems. At standard fluoroscopic exposures, FDXD performed in an equivalent

  19. SPECT detector system design based on embedded system

    International Nuclear Information System (INIS)

    Zhang Weizheng; Zhao Shujun; Zhang Lei; Sun Yuanling

    2007-01-01

    A single-photon emission computed tomography detector system based on embedded Linux designed. This system is composed of detector module, data acquisition module, ARM MPU module, network interface communication module and human machine interface module. Its software uses multithreading technology based on embedded Linux. It can achieve high speed data acquisition, real-time data correction and network data communication. It can accelerate the data acquisition and decrease the dead time. The accuracy and the stability of the system can be improved. (authors)

  20. A flat array large telescope concept for use on the moon, earth, and in space

    Science.gov (United States)

    Woodgate, Bruce E.

    1991-01-01

    An astronomical optical telescope concept is described which can provide very large collecting areas, of order 1000 sq m. This is an order of magnitude larger than the new generation of telescopes now being designed and built. Multiple gimballed flat mirrors direct the beams from a celestial source into a single telescope of the same aperture as each flat mirror. Multiple images of the same source are formed at the telescope focal plane. A beam combiner collects these images and superimposes them into a single image, onto a detector or spectrograph aperture. This telescope could be used on the earth, the moon, or in space.

  1. Full offline reconstruction in real-time with the LHCb detector

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00341115

    2016-01-01

    This document describes the novel, unique in High Energy Physics, real-time alignment and calibration of the full LHCb detector. The LHCb experiment has been designed as a dedicated heavy flavour physics experiment focused on the reconstruction of c and b hadrons. The LHCb detector is a single-arm forward spectrometer, which measures proton-proton interactions at the LHC. The operational bunch crossing rate is several orders of magnitude above the current abilities of data recording and storage. Therefore, a trigger system has been implemented to reduce this rate to an acceptable value. The LHCb trigger system has been redesigned during the 2013-2015 long shutdown, achieving oine-quality alignment and calibration online. It also allows analyses to be performed entirely at the trigger level. In addition, having the best performing reconstruction in the trigger gives the possibility to fully use the particle identification selection criteria and greatly increases the eciency, in particular for the selection of ...

  2. Evaluation of the dose distribution of dynamic conical conformal therapy using a C-arm mounted accelerator

    International Nuclear Information System (INIS)

    Nakagawa, Keiichi; Aoki, Yukimasa; Ohtomo, Kuni

    2001-01-01

    Conformal radiation therapy, which is widely utilized in Japan as a standard, highly precise technique has limited advantage in dose confinement because of its coplanar beam entry. An improved form of conformal therapy is delivered by a linac mounted on a C-arm rotatable gantry. The linac head was designed to move along the C-arm with a maximum angle of 60 degrees. Simultaneous rotation of the gantry creates a Dynamic Conical irradiation technique. Dynamic Conical Conformal Therapy (Dyconic Therapy) was developed by combining the technique with continuous MLC motion based on beam's eye views of the target volume. Dose distributions were measured in a phantom using film densitometry and compared with conventional conformal radiation therapy. The measurements showed that the dose distribution conformed to the target shape identified by CT. In addition, the dose distribution for a cancer patient was evaluated through the use of DVHs generated by a treatment planning system. These measurements showed that the dose distribution along the patient's long axis conformed to the shape of the target volume. DVH analysis, however, did not indicate superiority of the present technique over the conventional technique. Angulation of the C-arm gantry allowed the primary beam to strike a larger area of the therapy room. This necessitated adding shielding to the walls and ceiling of the treatment room. It was confirmed that the leakage radiation was reduced to a negligible level by adding an iron plate 20 cm thick to several places on the side walls, by adding an iron plate 9 cm thick to several places on the ceiling, and by increasing the thickness of the concrete ceiling from 70 to 140 cm. The possible usefulness of Dyconic Therapy was confirmed. (author)

  3. Application of focused-beam flat-sample method to synchrotron powder X-ray diffraction with anomalous scattering effect

    International Nuclear Information System (INIS)

    Tanaka, M; Katsuya, Y; Matsushita, Y

    2013-01-01

    The focused-beam flat-sample method (FFM), which is a method for high-resolution and rapid synchrotron X-ray powder diffraction measurements by combination of beam focusing optics, a flat shape sample and an area detector, was applied for diffraction experiments with anomalous scattering effect. The advantages of FFM for anomalous diffraction were absorption correction without approximation, rapid data collection by an area detector and good signal-to-noise ratio data by focusing optics. In the X-ray diffraction experiments of CoFe 2 O 4 and Fe 3 O 4 (By FFM) using X-rays near the Fe K absorption edge, the anomalous scattering effect between Fe/Co or Fe 2+ /Fe 3+ can be clearly detected, due to the change of diffraction intensity. The change of observed diffraction intensity as the incident X-ray energy was consistent with the calculation. The FFM is expected to be a method for anomalous powder diffraction.

  4. Rietveld analysis using powder diffraction data with anomalous scattering effect obtained by focused beam flat sample method

    Energy Technology Data Exchange (ETDEWEB)

    Tanaka, Masahiko, E-mail: masahiko@spring8.or.jp; Katsuya, Yoshio, E-mail: katsuya@spring8.or.jp; Sakata, Osami, E-mail: SAKATA.Osami@nims.go.jp [Synchrotron X-ray Station at SPring-8, National Institute for Materials Science 1-1-1 Kouto, Sayo, Hyogo 679-5198 (Japan)

    2016-07-27

    Focused-beam flat-sample method (FFM) is a new trial for synchrotron powder diffraction method, which is a combination of beam focusing optics, flat shape powder sample and area detectors. The method has advantages for X-ray diffraction experiments applying anomalous scattering effect (anomalous diffraction), because of 1. Absorption correction without approximation, 2. High intensity X-rays of focused incident beams and high signal noise ratio of diffracted X-rays 3. Rapid data collection with area detectors. We applied the FFM to anomalous diffraction experiments and collected synchrotron X-ray powder diffraction data of CoFe{sub 2}O{sub 4} (inverse spinel structure) using X-rays near Fe K absorption edge, which can distinguish Co and Fe by anomalous scattering effect. We conducted Rietveld analyses with the obtained powder diffraction data and successfully determined the distribution of Co and Fe ions in CoFe{sub 2}O{sub 4} crystal structure.

  5. Development of a simple detector response function generation program: The CEARDRFs code

    Energy Technology Data Exchange (ETDEWEB)

    Wang Jiaxin, E-mail: jwang3@ncsu.edu [Center for Engineering Applications of Radioisotopes (CEAR), Department of Nuclear Engineering, North Carolina State University, Raleigh, NC 27695 (United States); Wang Zhijian; Peeples, Johanna [Center for Engineering Applications of Radioisotopes (CEAR), Department of Nuclear Engineering, North Carolina State University, Raleigh, NC 27695 (United States); Yu Huawei [Center for Engineering Applications of Radioisotopes (CEAR), Department of Nuclear Engineering, North Carolina State University, Raleigh, NC 27695 (United States); College of Geo-Resources and Information, China University of Petroleum, Qingdao, Shandong 266555 (China); Gardner, Robin P. [Center for Engineering Applications of Radioisotopes (CEAR), Department of Nuclear Engineering, North Carolina State University, Raleigh, NC 27695 (United States)

    2012-07-15

    A simple Monte Carlo program named CEARDRFs has been developed to generate very accurate detector response functions (DRFs) for scintillation detectors. It utilizes relatively rigorous gamma-ray transport with simple electron transport, and accounts for two phenomena that have rarely been treated: scintillator non-linearity and the variable flat continuum part of the DRF. It has been proven that these physics and treatments work well for 3 Multiplication-Sign 3 Double-Prime and 6 Multiplication-Sign 6 Double-Prime cylindrical NaI detector in CEAR's previous work. Now this approach has been expanded to cover more scintillation detectors with various common shapes and sizes. Benchmark experiments of 2 Multiplication-Sign 2 Double-Prime cylindrical BGO detector and 2 Multiplication-Sign 4 Multiplication-Sign 16 Double-Prime rectangular NaI detector have been carried out at CEAR with various radiactive sources. The simulation results of CEARDRFs have also been compared with MCNP5 calculations. The benchmark and comparison show that CEARDRFs can generate very accurate DRFs (more accurate than MCNP5) at a very fast speed (hundred times faster than MCNP5). The use of this program can significantly increase the accuracy of applications relying on detector spectroscopy like prompt gamma-ray neutron activation analysis, X-ray fluorescence analysis, oil well logging and homeland security. - Highlights: Black-Right-Pointing-Pointer CEARDRF has been developed to generate detector response functions (DRFs) for scintillation detectors a. Black-Right-Pointing-Pointer Generated DRFs are very accurate. Black-Right-Pointing-Pointer Simulation speed is hundreds of times faster than MCNP5. Black-Right-Pointing-Pointer It utilizes rigorous gamma-ray transport with simple electron transport. Black-Right-Pointing-Pointer It also accounts for scintillator non-linearity and the variable flat continuum part.

  6. Online C-arm calibration using a marked guide wire for 3D reconstruction of pulmonary arteries

    Science.gov (United States)

    Vachon, Étienne; Miró, Joaquim; Duong, Luc

    2017-03-01

    3D reconstruction of vessels from 2D X-ray angiography is highly relevant to improve the visualization and the assessment of vascular structures such as pulmonary arteries by interventional cardiologists. However, to ensure a robust and accurate reconstruction, C-arm gantry parameters must be properly calibrated to provide clinically acceptable results. Calibration procedures often rely on calibration objects and complex protocol which is not adapted to an intervention context. In this study, a novel calibration algorithm for C-arm gantry is presented using the instrumentation such as catheters and guide wire. This ensures the availability of a minimum set of correspondences and implies minimal changes to the clinical workflow. The method was evaluated on simulated data and on retrospective patient datasets. Experimental results on simulated datasets demonstrate a calibration that allows a 3D reconstruction of the guide wire up to a geometric transformation. Experiments with patients datasets show a significant decrease of the retro projection error to 0.17 mm 2D RMS. Consequently, such procedure might contribute to identify any calibration drift during the intervention.

  7. Device including a contact detector

    DEFF Research Database (Denmark)

    2011-01-01

    arms (12) may extend from the supporting body in co-planar relationship with the first surface. The plurality of cantilever arms (12) may extend substantially parallel to each other and each of the plurality of cantilever arms (12) may include an electrical conductive tip for contacting the area......The present invention relates to a probe for determining an electrical property of an area of a surface of a test sample, the probe is intended to be in a specific orientation relative to the test sample. The probe may comprise a supporting body defining a first surface. A plurality of cantilever...... of the test sample by movement of the probe relative to the surface of the test sample into the specific orientation.; The probe may further comprise a contact detector (14) extending from the supporting body arranged so as to contact the surface of the test sample prior to any one of the plurality...

  8. Radiation dose reduction and new image modalities development for interventional C-arm imaging system

    Science.gov (United States)

    Niu, Kai

    Cardiovascular disease and stroke are the leading health problems and causes of death in the US. Due to the minimally invasive nature of the evolution of image guided techniques, interventional radiological procedures are becoming more common and are preferred in treating many cardiovascular diseases and strokes. In addition, with the recent advances in hardware and device technology, the speed and efficacy of interventional treatment has significantly improved. This implies that more image modalities can be developed based on the current C-arm system and patients treated in interventional suites can potentially experience better health outcomes. However, during the treatment patients are irradiated with substantial amounts of ionizing radiation with a high dose rate (digital subtraction angiography (DSA) with 3muGy/frame and 3D cone beam CT image with 0.36muGy/frame for a Siemens Artis Zee biplane system) and/or a long irradiation time (a roadmapping image sequence can be as long as one hour during aneurysm embolization). As a result, the patient entrance dose is extremely high. Despite the fact that the radiation dose is already substantial, image quality is not always satisfactory. By default a temporal average is used in roadmapping images to overcome poor image quality, but this technique can result in motion blurred images. Therefore, reducing radiation dose while maintaining or even improving the image quality is an important area for continued research. This thesis is focused on improving the clinical applications of C-arm cone beam CT systems in two ways: (1) Improve the performance of current image modalities on the C-arm system. (2) Develop new image modalities based on the current system. To be more specific, the objectives are to reduce radiation dose for current modalities (e.g., DSA, fluoroscopy, roadmapping, and cone beam CT) and enable cone beam CT perfusion and time resolved cone beam CT angiography that can be used to diagnose and triage acute

  9. MO-DE-207A-06: ECG-Gated CT Reconstruction for a C-Arm Inverse Geometry X-Ray System

    Energy Technology Data Exchange (ETDEWEB)

    Slagowski, JM; Dunkerley, DAP [MA Speidel, University of Wisconsin - Madison, Madison, WI (United States)

    2016-06-15

    Purpose: To obtain ECG-gated CT images from truncated projection data acquired with a C-arm based inverse geometry fluoroscopy system, for the purpose of cardiac chamber mapping in interventional procedures. Methods: Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system with a scanned multisource x-ray tube and a photon-counting detector mounted to a C-arm. In the proposed method, SBDX short-scan rotational acquisition is performed followed by inverse geometry CT (IGCT) reconstruction and segmentation of contrast-enhanced objects. The prior image constrained compressed sensing (PICCS) framework was adapted for IGCT reconstruction to mitigate artifacts arising from data truncation and angular undersampling due to cardiac gating. The performance of the reconstruction algorithm was evaluated in numerical simulations of truncated and non-truncated thorax phantoms containing a dynamic ellipsoid to represent a moving cardiac chamber. The eccentricity of the ellipsoid was varied at frequencies from 1–1.5 Hz. Projection data were retrospectively sorted into 13 cardiac phases. Each phase was reconstructed using IGCT-PICCS, with a nongated gridded FBP (gFBP) prior image. Surface accuracy was determined using Dice similarity coefficient and a histogram of the point distances between the segmented surface and ground truth surface. Results: The gated IGCT-PICCS algorithm improved surface accuracy and reduced streaking and truncation artifacts when compared to nongated gFBP. For the non-truncated thorax with 1.25 Hz motion, 99% of segmented surface points were within 0.3 mm of the 15 mm diameter ground truth ellipse, versus 1.0 mm for gFBP. For the truncated thorax phantom with a 40 mm diameter ellipse, IGCT-PICCS surface accuracy measured 0.3 mm versus 7.8 mm for gFBP. Dice similarity coefficient was 0.99–1.00 (IGCT-PICCS) versus 0.63–0.75 (gFBP) for intensity-based segmentation thresholds ranging from 25–75% maximum contrast. Conclusions: The

  10. An Optimized Spline-Based Registration of a 3D CT to a Set of C-Arm Images

    Directory of Open Access Journals (Sweden)

    2006-01-01

    Full Text Available We have developed an algorithm for the rigid-body registration of a CT volume to a set of C-arm images. The algorithm uses a gradient-based iterative minimization of a least-squares measure of dissimilarity between the C-arm images and projections of the CT volume. To compute projections, we use a novel method for fast integration of the volume along rays. To improve robustness and speed, we take advantage of a coarse-to-fine processing of the volume/image pyramids. To compute the projections of the volume, the gradient of the dissimilarity measure, and the multiresolution data pyramids, we use a continuous image/volume model based on cubic B-splines, which ensures a high interpolation accuracy and a gradient of the dissimilarity measure that is well defined everywhere. We show the performance of our algorithm on a human spine phantom, where the true alignment is determined using a set of fiducial markers.

  11. Evaluation of malrotation following intramedullary nailing in a femoral shaft fracture model: Can a 3D c-arm improve accuracy?

    Science.gov (United States)

    Ramme, Austin J; Egol, Jonathan; Chang, Gregory; Davidovitch, Roy I; Konda, Sanjit

    2017-07-01

    Difficulty determining anatomic rotation following intramedullary (IM) nailing of the femur continues to be problematic for surgeons. Clinical exam and fluoroscopic imaging of the hip and knee have been used to estimate femoral version, but are inaccurate. We hypothesize that 3D c-arm imaging can be used to accurately measure femoral version following IM nailing of femur fractures to prevent rotational malreduction. A midshaft osteotomy was created in a femur Sawbone to simulate a transverse diaphyseal fracture. An intramedullary (IM) nail was inserted into the Sawbone femur without locking screws or cephalomedullary fixation. A goniometer was used to simulate four femoral version situations after IM nailing: 20° retroversion, 0° version, 15° anteversion, and 30° anteversion. In each simulated position, 3D c-arm imaging and, for comparison purposes, perfect lateral radiographs of the knee and hip were performed. The femoral version of each simulated 3D and fluoroscopic case was measured and the results were tabulated. The measured version from the 3D c-arm images was 22.25° retroversion, 0.66° anteversion, 19.53° anteversion, and 25.15° anteversion for the simulated cases of 20° retroversion, 0° version, 15° anteversion, and 30° anteversion, respectively. The lateral fluoroscopic views were measured to be 9.66° retroversion, 12.12° anteversion, 20.91° anteversion, and 18.77° anteversion for the simulated cases, respectively. This study demonstrates the utility of a novel intraoperative method to evaluate femur rotational malreduction following IM nailing. The use of 3D c-arm imaging to measure femoral version offers accuracy and reproducibility. Copyright © 2017 Elsevier Ltd. All rights reserved.

  12. Counterbalanced radiation detection system

    International Nuclear Information System (INIS)

    Platz, W.

    1987-01-01

    A counterbalanced radiation detection system is described comprising: (a) a stand; (b) a first radiation detector; (c) a first radiation detector arm means for tiltably connecting the first radiation detector with the stand; (d) a second radiation detector; (e) a second radiation detector arm means for tiltably connecting the second radiation detector with the stand, whereby the tilting angles of the radiation detector arm means define a distance between the radiation detectors; and (f) a torque transforming means connected between the first and second radiation detector arm means for transforming the torque created by one of the radiation detectors in a sense opposed to the torque created by the other radiation detector

  13. Advanced radiation detector development: Advanced semiconductor detector development: Development of a oom-temperature, gamma ray detector using gallium arsenide to develop an electrode detector

    International Nuclear Information System (INIS)

    Knoll, G.F.

    1995-11-01

    The advanced detector development project at the University of Michigan has completed the first full year of its current funding. Our general goals are the development of radiation detectors and spectrometers that are capable of portable room temperature operation. Over the past 12 months, we have worked primarily in the development of semiconductor spectrometers with open-quotes single carrierclose quotes response that offer the promise of room temperature operation and good energy resolution in gamma ray spectroscopy. We have also begun a small scale effort at investigating the properties of a small non-spectroscopic detector system with directional characteristics that will allow identification of the approximate direction in which gamma rays are incident. These activities have made use of the extensive clean room facilities at the University of Michigan for semiconductor device fabrication, and also the radiation measurement capabilities provided in our laboratory in the Phoenix Building on the North Campus. In addition to our laboratory based activities, Professor Knoll has also been a participant in several Department of Energy review activities held in the Forrestal Building and at the Germantown site. The most recent of these has been service on a DOE review panel chaired by Dr. Hap Lamonds that is reviewing the detector development programs supported through the Office of Arms Control and International Security

  14. Interesting detector shapes for 3rd generation CT

    Energy Technology Data Exchange (ETDEWEB)

    Kachelriess, Marc [Erlangen-Nuernberg Univ., Erlangen (Germany). Inst. of Medical Physics (IMP)

    2011-07-01

    Third generation CT scanners typically comprise detectors which are flat or whose shape is the segment of a cylinder or a sphere that is focussed onto the focal spot of the X-ray source. There appear to be two design criteria that favor this choice of detector shape. One is the possibility of performing fan-beam and cone-beam filtered backprojection in the native geometry (without rebinning) and the other criterium is to enable the use of focussed anti scatter grids. It is less known, however, that other detector shapes may also have these properties. While these designs have been evaluated for 2D CT from a rather theoretical standpoint more than one decade ago we revisit and generalize these considerations, extend them to 3D circular, sequential and spiral cone-beam CT and propose an optimal design in terms of detector costs. (orig.)

  15. C-arm cone-beam CT combined with a new electromagnetic navigation system for guidance of percutaneous needle biopsies. Initial clinical experience

    Energy Technology Data Exchange (ETDEWEB)

    Kickuth, R.; Reichling, C.; Bley, T.; Hahn, D.; Ritter, C. [University Hospital of Wuerzburg (Germany). Inst. of Diagnostic and Interventional Radiology

    2015-07-15

    To evaluate the feasibility and efficacy of C-arm fluoroscopic cone-beam computed tomography (CACT) in combination with a new electromagnetic tracking (EMT) system for needle guidance during percutaneous biopsies. 53 patients were referred for biopsy of thoracic (n = 19) and abdominal (n = 34) lesions. CT-like images of the anatomical region of interest (ROI) were generated using a flat panel-based angiographic system. These images were transmitted to an EMT system. A coaxial puncture needle with a sensor in its tip was connected with the navigation system and tracked into an electromagnetic field created via a field generator. Data generated within this field were merged with the CACT images. On a monitor both the anatomical ROI and needle tip position were displayed to enable precise needle insertion into the target. Through the coaxial needle, biopsy specimens for the histologic evaluation were extracted. Number of representative biopsy samples, number of core biopsies/patient, total procedure time, dose-area product, fluoroscopic time, and complications were recorded. 53 CACT/EMT-guided biopsy procedures were performed, 48 of which (91 %) yielded representative tissue samples. Four core biopsies were obtained from each patient. 40 (75 %) lesions were malignant and 13 (25 %) lesions were benign. The total procedure time was 9 ± 5 min (range, 3 - 23 min), fluoroscopic time was 0.8 ± 0.4 min (range, 0.4 - 2 min). The mean dose-area product (cGy cm{sup 2}) was 7373 (range, 895 - 26 904). The rate of complications (1 pneumothorax, 2 hemoptyses) was 6 %. CACT combined with EMT appears to be a feasible and effective technique for the guidance of percutaneous biopsies with a low rate of therapeutically relevant complications.

  16. C-arm cone-beam CT combined with a new electromagnetic navigation system for guidance of percutaneous needle biopsies. Initial clinical experience

    International Nuclear Information System (INIS)

    Kickuth, R.; Reichling, C.; Bley, T.; Hahn, D.; Ritter, C.

    2015-01-01

    To evaluate the feasibility and efficacy of C-arm fluoroscopic cone-beam computed tomography (CACT) in combination with a new electromagnetic tracking (EMT) system for needle guidance during percutaneous biopsies. 53 patients were referred for biopsy of thoracic (n = 19) and abdominal (n = 34) lesions. CT-like images of the anatomical region of interest (ROI) were generated using a flat panel-based angiographic system. These images were transmitted to an EMT system. A coaxial puncture needle with a sensor in its tip was connected with the navigation system and tracked into an electromagnetic field created via a field generator. Data generated within this field were merged with the CACT images. On a monitor both the anatomical ROI and needle tip position were displayed to enable precise needle insertion into the target. Through the coaxial needle, biopsy specimens for the histologic evaluation were extracted. Number of representative biopsy samples, number of core biopsies/patient, total procedure time, dose-area product, fluoroscopic time, and complications were recorded. 53 CACT/EMT-guided biopsy procedures were performed, 48 of which (91 %) yielded representative tissue samples. Four core biopsies were obtained from each patient. 40 (75 %) lesions were malignant and 13 (25 %) lesions were benign. The total procedure time was 9 ± 5 min (range, 3 - 23 min), fluoroscopic time was 0.8 ± 0.4 min (range, 0.4 - 2 min). The mean dose-area product (cGy cm 2 ) was 7373 (range, 895 - 26 904). The rate of complications (1 pneumothorax, 2 hemoptyses) was 6 %. CACT combined with EMT appears to be a feasible and effective technique for the guidance of percutaneous biopsies with a low rate of therapeutically relevant complications.

  17. Rapid learning of magnetic compass direction by C57BL/6 mice in a 4-armed 'plus' water maze.

    Science.gov (United States)

    Phillips, John B; Youmans, Paul W; Muheim, Rachel; Sloan, Kelly A; Landler, Lukas; Painter, Michael S; Anderson, Christopher R

    2013-01-01

    Magnetoreception has been demonstrated in all five vertebrate classes. In rodents, nest building experiments have shown the use of magnetic cues by two families of molerats, Siberian hamsters and C57BL/6 mice. However, assays widely used to study rodent spatial cognition (e.g. water maze, radial arm maze) have failed to provide evidence for the use of magnetic cues. Here we show that C57BL/6 mice can learn the magnetic direction of a submerged platform in a 4-armed (plus) water maze. Naïve mice were given two brief training trials. In each trial, a mouse was confined to one arm of the maze with the submerged platform at the outer end in a predetermined alignment relative to magnetic north. Between trials, the training arm and magnetic field were rotated by 180(°) so that the mouse had to swim in the same magnetic direction to reach the submerged platform. The directional preference of each mouse was tested once in one of four magnetic field alignments by releasing it at the center of the maze with access to all four arms. Equal numbers of responses were obtained from mice tested in the four symmetrical magnetic field alignments. Findings show that two training trials are sufficient for mice to learn the magnetic direction of the submerged platform in a plus water maze. The success of these experiments may be explained by: (1) absence of alternative directional cues (2), rotation of magnetic field alignment, and (3) electromagnetic shielding to minimize radio frequency interference that has been shown to interfere with magnetic compass orientation of birds. These findings confirm that mice have a well-developed magnetic compass, and give further impetus to the question of whether epigeic rodents (e.g., mice and rats) have a photoreceptor-based magnetic compass similar to that found in amphibians and migratory birds.

  18. Studies of frequency dependent C-V characteristics of neutron irradiated p+-n silicon detectors

    International Nuclear Information System (INIS)

    Li, Zheng; Kraner, H.W.

    1990-10-01

    Frequency-dependent capacitance-voltage fluence (C-V) characteristics of neutron irradiated high resistivity silicon p + -n detectors have been observed up to a fluence of 8.0 x 10 12 n/cm 2 . It has been found that frequency dependence of the deviation of the C-V characteristic (from its normal V -1/2 dependence), is strongly dependent on the ratio of the defect density and the effective doping density N t /N' d . As the defect density approaches the effective dopant density, or N t /N' d → 1, the junction capacitance eventually assumes the value of the detector geometry capacitance at high frequencies (f ≤ 10 5 Hz), independent of voltage. A two-trap-level model using the concept of quasi-fermi levels has been developed, which predicts both the effects of C-V frequency dependence and dopant compensation observed in this study

  19. Unitarity in three-dimensional flat space higher spin theories

    International Nuclear Information System (INIS)

    Grumiller, D.; Riegler, M.; Rosseel, J.

    2014-01-01

    We investigate generic flat-space higher spin theories in three dimensions and find a no-go result, given certain assumptions that we spell out. Namely, it is only possible to have at most two out of the following three properties: unitarity, flat space, non-trivial higher spin states. Interestingly, unitarity provides an (algebra-dependent) upper bound on the central charge, like c=42 for the Galilean W_4"("2"−"1"−"1") algebra. We extend this no-go result to rule out unitary “multi-graviton” theories in flat space. We also provide an example circumventing the no-go result: Vasiliev-type flat space higher spin theory based on hs(1) can be unitary and simultaneously allow for non-trivial higher-spin states in the dual field theory.

  20. R&D studies of a RICH detector using pressurized C$_{4}$F$_{8}$O radiator gas and a CsI-based gaseous photon detector

    CERN Document Server

    Agócs, A.G; Barnaföldi, G.G; Bellwied, R; Bencédi, G; Bencze, G; Berényi, D; Boldizsár, L; Chattopadhyay, S; Chinellato, D.D; Cindolo, F; Das-Bose, L; Das, D; Das, K; De Cataldo, G; Di Bari, D; Di Mauro, A; Futó, E; Garcia, E; Hamar, G; Harton, A; Jimenez, R.T; Kim, D.W; Kim, J.S; Knospe, A; Kovacs, L; Lévai, P; Markert, C; Martinengo, P; Molnar, L; Nappi, E; Olah, L; Paic, G; Pastore, C; Patino, M.E; Peskov, V; Pinsky, L; Piuz, F; Pochybova, S; Sgura, I; Sinha, T; Song, J; Timmins, A; Van Beelen, J.B; Varga, D; Volpe, G; Weber, M; Xaplanteris, L; Yi, J; Yoo, I.-K

    2013-01-01

    We report on studies of layout and performance of a new Ring Imaging Cherenkov detector using for the fi rst time pressurized C 4 F 8 O radiator gas and a photon detector consisting of a MWPC equipped with a CsI photocathode. In particular, we present here the results of beam tests of a MWPC having an adjustable anode – cathode gap, aiming at the optimization of single photoelectron detection and Cherenkov angle resolution. This system was proposed as a Very High Momentum Particle Identi fi cation (VHMPID) upgrade for the ALICE experiment at LHC to provide charged hadron track-by-track identi fi cation in the momentum range 5 – 25 GeV/c.

  1. Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams

    DEFF Research Database (Denmark)

    Azangwe, Godfrey; Grochowska, Paulina; Georg, Dietmar

    2014-01-01

    -doped aluminium oxide (Al2O3:C), organic plastic scintillators, diamond detectors, liquid filled ion chamber, and a range of small volume air filled ionization chambers (volumes ranging from 0.002 cm3 to 0.3 cm3). All detector measurements were corrected for volume averaging effect and compared with dose ratios...... measurements, the authors recommend the use of detectors that require relatively little correction, such as unshielded diodes, diamond detectors or microchambers, and solid state detectors such as alanine, TLD, Al2O3:C, or scintillators....

  2. Gamma ray and neutrino detector facility (GRANDE), Task C

    International Nuclear Information System (INIS)

    Sobel, H.W.; Yodh, G.B.

    1991-08-01

    GRANDE is an imaging, water Cerenkov detector, which combines in one facility an extensive air shower array and a high-energy neutrino detector. We proposed that the detector be constructed in phases, beginning with an active detector area of 31,000 m 2 (GRANDE-I) 2 and expanding to a final size of 100,000--150,00 m 2 . Some of the characteristics of GRANDE-I are discussed in this paper

  3. Sensitization by UV light of α-Al{sub 2}O{sub 3}:C polycrystalline detectors

    Energy Technology Data Exchange (ETDEWEB)

    Meira B, L. C.; Rubio F, H.; Neres de A, E. [Centro de Desenvolvimento da Tecnologia Nuclear - CNEN, Laboratorio de Dosimetria Termoluminicente, Av. Antonio Carlos 6627, Campus UFMG, CEP 31270-901, Belo Horizonte (Brazil); Santos, A., E-mail: hrf@cdtn.br [Centro de Desenvolvimento da Tecnologia Nuclear - CNEN, Laboratorio de Microesferas Gel, Av. Antonio Carlos 6627, Campus UFMG, CEP 31270-901, Belo Horizonte (Brazil)

    2014-08-15

    This paper describes an increase in sensitivity to gamma and beta radiation on α-Al{sub 2}O{sub 3}:C polycrystalline detector, which has been produced by a sol-gel process, following previous exposure to ultraviolet light. The increased sensitivity of the detector as a function of the exposure time and ultraviolet wavelength was studied. Since the main luminescent centers have emission peaks at different wavelengths, selective measurements of thermoluminescent emission intensity were done, in order to investigate the possible conversion of centers as a result of the exposition to ultraviolet light. Experimental results indicate that the nature and parameters of the luminescent centers in α-Al{sub 2}O{sub 3}:C sol-gel material can be very different of those in α-Al{sub 2}O{sub 3}:C single crystal. (author)

  4. Physics prospects and technical status of ATLAS Forward proton detectors

    CERN Document Server

    Giacobbe, Benedetto; The ATLAS collaboration

    2017-01-01

    The aim of the ATLAS Forward Proton (AFP) detector system is the measurement of protons scattered diffractively or electromagnetically at very small angles. The first arm of the system was installed last year and AFP took data in several commissioning and physics runs. The installation of the second arm is ongoing and will be completed in time for the 2017 data taking period. This will allow measurements of processes with two forward protons: central diffraction, exclusive production, and two-photon processes. The presentation will cover the physics programme, the early the experience from the first year of running, the status of the second-arm installation and the data taking plans.

  5. Intraprocedural blood volume measurement using C-arm CT as a predictor for treatment response of malignant liver tumours undergoing repetitive transarterial chemoembolization (TACE)

    International Nuclear Information System (INIS)

    Vogl, Thomas J.; Schaefer, Patrik; Lehnert, Thomas; Mbalisike, Emmanuel; Hammerstingl, Renate; Eichler, Katrin; Zangos, Stephan; Nour-Eldin, Nour-Eldin A.; Ackermann, Hanns; Naguib, Nagy N.N.

    2016-01-01

    To evaluate feasibility of measuring parenchymal blood volume (PBV) of malignant hepatic tumours using C-arm CT, test the changes in PBV following repeated transarterial chemoembolization (TACE) and correlate these changes with the change in tumour size in MRI. 111 patients with liver malignancy were included. Patients underwent MRI and TACE in a 4- to 6-week interval. During intervention C-arm CT was performed. Images were post-processed to generate PBV maps. Blood volume data in C-arm CT and change in size in MRI were evaluated. The correlation between PBV and size was tested using Spearman rank test. Pre-interventional PBV maps showed a mean blood volume of 84.5 ml/1000 ml ± 62.0, follow-up PBV maps after multiple TACE demonstrated 61.1 ml/1000 ml ± 57.5. The change in PBV was statistically significant (p = 0.02). Patients with initial tumour blood volume >100 ml/1000 ml dropped 7.1 % in size and 47.2 % in blood volume; 50-100 ml/1000 ml dropped 4.6 % in size and 25.7 % in blood volume; and <50 ml/1000 ml decreased 2.8 % in size and increased 82.2 % in blood volume. PBV measurement of malignant liver tumours using C-arm CT is feasible. Following TACE PBV decreased significantly. Patients with low initial PBV show low local response rates and further increase in blood volume, whereas high initial tumour PBV showed better response to TACE. (orig.)

  6. Upper ankle joint space detection on low contrast intraoperative fluoroscopic C-arm projections

    Science.gov (United States)

    Thomas, Sarina; Schnetzke, Marc; Brehler, Michael; Swartman, Benedict; Vetter, Sven; Franke, Jochen; Grützner, Paul A.; Meinzer, Hans-Peter; Nolden, Marco

    2017-03-01

    Intraoperative mobile C-arm fluoroscopy is widely used for interventional verification in trauma surgery, high flexibility combined with low cost being the main advantages of the method. However, the lack of global device-to- patient orientation is challenging, when comparing the acquired data to other intrapatient datasets. In upper ankle joint fracture reduction accompanied with an unstable syndesmosis, a comparison to the unfractured contralateral site is helpful for verification of the reduction result. To reduce dose and operation time, our approach aims at the comparison of single projections of the unfractured ankle with volumetric images of the reduced fracture. For precise assessment, a pre-alignment of both datasets is a crucial step. We propose a contour extraction pipeline to estimate the joint space location for a prealignment of fluoroscopic C-arm projections containing the upper ankle joint. A quadtree-based hierarchical variance comparison extracts potential feature points and a Hough transform is applied to identify bone shaft lines together with the tibiotalar joint space. By using this information we can define the coarse orientation of the projections independent from the ankle pose during acquisition in order to align those images to the volume of the fractured ankle. The proposed method was evaluated on thirteen cadaveric datasets consisting of 100 projections each with manually adjusted image planes by three trauma surgeons. The results show that the method can be used to detect the joint space orientation. The correlation between angle deviation and anatomical projection direction gives valuable input on the acquisition direction for future clinical experiments.

  7. Direct navigation on 3D rotational x-ray data acquired with a mobile propeller C-arm: accuracy and application in functional endoscopic sinus surgery

    International Nuclear Information System (INIS)

    Kraats, Everine B van de; Carelsen, Bart; Fokkens, Wytske J; Boon, Sjirk N; Noordhoek, Niels; Niessen, Wiro J; Walsum, Theo van

    2005-01-01

    Recently, three-dimensional (3D) rotational x-ray imaging has been combined with navigation technology, enabling direct 3D navigation for minimally invasive image guided interventions. In this study, phantom experiments are used to determine the accuracy of such a navigation set-up for a mobile C-arm with propeller motion. After calibration of the C-arm system, the accuracy is evaluated by pinpointing divots on a special-purpose phantom with known geometry. This evaluation is performed both with and without C-arm motion in between calibration and registration for navigation. The variation caused by each of the individual transformations in the calibration and registration process is also studied. The feasibility of direct navigation on 3D rotational x-ray images for functional endoscopic sinus surgery has been evaluated in a cadaver navigation experiment. Navigation accuracy was approximately 1.0 mm, which is sufficient for functional endoscopic sinus surgery. C-arm motion in between calibration and registration slightly degraded the registration accuracy by approximately 0.3 mm. Standard deviations of each of the transformations were in the range 0.15-0.31 mm. In the cadaver experiment, the navigation images were considered in good correspondence with the endoscopic images by an experienced ENT surgeon. Availability of 3D localization information provided by the navigation system was considered valuable by the ENT surgeon

  8. Calculation of bulk charge and electric field profiles in one-open-face coaxial γ-detectors using experimental C-U characteristics

    International Nuclear Information System (INIS)

    Kirnas, I.G.; Litovchenko, P.G.; Petrosyan, E.E.; Pashchuk, N.N.

    1987-01-01

    Experimental C-U characteristics of one-open-face coaxial detectors are employed in deriving an analytic expression that describes the distributions of ionized impurity bulk charge e(N a - N d ) and electric field E(r) for arbitrary variations of N a - N d in the detector volume. As an example, e(N a - N d ) = f(r) and E(r) are calculated for a Ge(Li)-detector whose experimental C-U characteristics is approximated by a power law with exponent two. (author)

  9. Single-layer and dual-layer contrast-enhanced mammography using amorphous selenium flat panel detectors

    Energy Technology Data Exchange (ETDEWEB)

    Allec, N; Abbaszadeh, S; Karim, K S, E-mail: nallec@uwaterloo.ca [Department of Electrical and Computer Engineering, University of Waterloo, 200 University Avenue West, Waterloo N2L 3G1 (Canada)

    2011-09-21

    The accumulation of injected contrast agents allows the image enhancement of lesions through the use of contrast-enhanced mammography. In this technique, the combination of two acquired images is used to create an enhanced image. There exist several methods to acquire the images to be combined, which include dual energy subtraction using a single detection layer that suffers from motion artifacts due to patient motion between image acquisition. To mitigate motion artifacts, a detector composed of two layers may be used to simultaneously acquire the low and high energy images. In this work, we evaluate both of these methods using amorphous selenium as the detection material to find the system parameters (tube voltage, filtration, photoconductor thickness and relative intensity ratio) leading to the optimal performance. We then compare the performance of the two detectors under the variation of contrast agent concentration, tumor size and dose. The detectability was found to be most comparable at the lower end of the evaluated factors. The single-layer detector not only led to better contrast, due to its greater spectral separation capabilities, but also had lower quantum noise. The single-layer detector was found to have a greater detectability by a factor of 2.4 for a 2.5 mm radius tumor having a contrast agent concentration of 1.5 mg ml{sup -1} in a 4.5 cm thick 50% glandular breast. The inclusion of motion artifacts in the comparison is part of ongoing research efforts.

  10. Single-layer and dual-layer contrast-enhanced mammography using amorphous selenium flat panel detectors

    Science.gov (United States)

    Allec, N.; Abbaszadeh, S.; Karim, K. S.

    2011-09-01

    The accumulation of injected contrast agents allows the image enhancement of lesions through the use of contrast-enhanced mammography. In this technique, the combination of two acquired images is used to create an enhanced image. There exist several methods to acquire the images to be combined, which include dual energy subtraction using a single detection layer that suffers from motion artifacts due to patient motion between image acquisition. To mitigate motion artifacts, a detector composed of two layers may be used to simultaneously acquire the low and high energy images. In this work, we evaluate both of these methods using amorphous selenium as the detection material to find the system parameters (tube voltage, filtration, photoconductor thickness and relative intensity ratio) leading to the optimal performance. We then compare the performance of the two detectors under the variation of contrast agent concentration, tumor size and dose. The detectability was found to be most comparable at the lower end of the evaluated factors. The single-layer detector not only led to better contrast, due to its greater spectral separation capabilities, but also had lower quantum noise. The single-layer detector was found to have a greater detectability by a factor of 2.4 for a 2.5 mm radius tumor having a contrast agent concentration of 1.5 mg ml-1 in a 4.5 cm thick 50% glandular breast. The inclusion of motion artifacts in the comparison is part of ongoing research efforts.

  11. The L3+C detector, a unique tool-set to study cosmic rays

    International Nuclear Information System (INIS)

    Adriani, O.; Akker, M. van den; Banerjee, S.; Baehr, J.; Betev, B.; Bourilkov, D.; Bottai, S.; Bobbink, G.; Cartacci, A.; Chemarin, M.; Chen, G.; Chen, H.S.; Chiarusi, T.; Dai, C.J.; Ding, L.K.; Duran, I.; Faber, G.; Fay, J.; Grabosch, H.J.; Groenstege, H.; Guo, Y.N.; Gupta, S.; Haller, Ch.; Hayashi, Y.; He, Z.X.; Hebbeker, T.; Hofer, H.; Hoferjun, H.; Huo, A.X.; Ito, N.; Jing, C.L.; Jones, L.; Kantserov, V.; Kawakami, S.; Kittel, W.; Koenig, A.C.; Kok, E.; Korn, A.; Kuang, H.H.; Kuijpers, J.; Ladron de Guevara, P.; Le Coultre, P.; Lei, Y.; Leich, H.; Leiste, R.; Li, D.; Li, L.; Li, Z.C.; Liu, Z.A.; Liu, H.T.; Lohmann, W.; Lu, Y.S.; Ma, X.H.; Ma, Y.Q.; Mil, A. van; Monteleoni, B.; Nahnhauer, R.; Pauss, F.; Parriaud, J.-F.; Petersen, B.; Pohl, M.; Qing, C.R.; Ramelli, R.; Ravindran, K.C.; Rewiersma, P.; Rojkov, A.; Saidi, R.; Schmitt, V.; Schoeneich, B.; Schotanus, D.J.; Shen, C.Q.; Sulanke, H.; Tang, X.W.; Timmermans, C.; Tonwar, S.; Trowitzsch, G.; Unger, M.; Verkooijen, H.; Wang, X.L.; Wang, X.W.; Wang, Z.M.; Wijk, R. van; Wijnen, Th.A.M.; Wilkens, H.; Xu, Y.P.; Xu, Z.Z.; Yang, C.G.; Yang, X.F.; Yao, Z.G.; Yu, Z.Q.; Zhang, S.; Zhu, G.Y.; Zhu, Q.Q.; Zhuang, H.L.; Zwart, A.N.M.

    2002-01-01

    The L3 detector at the CERN electron-positron collider, LEP, has been employed for the study of cosmic ray muons. The muon spectrometer of L3 consists of a set of high-precision drift chambers installed inside a magnet with a volume of about 1000 m 3 and a field of 0.5 T. Muon momenta are measured with a resolution of a few percent at 50 GeV. The detector is located under 30 m of overburden. A scintillator air shower array of 54 m by 30 m is installed on the roof of the surface hall above L3 in order to estimate the energy and the core position of the shower associated with a sample of detected muons. Thanks to the unique properties of the L3+C detector, muon research topics relevant to various current problems in cosmic ray and particle astrophysics can be studied

  12. The L3+C detector, a unique tool-set to study cosmic rays

    CERN Document Server

    Adriani, O; Banerjee, S; Bähr, J; Betev, B L; Bourilkov, D; Bottai, S; Bobbink, Gerjan J; Cartacci, A M; Chemarin, M; Chen, G; Chen He Sheng; Chiarusi, T; Dai Chang Jiang; Ding, L K

    2002-01-01

    The L3 detector at the CERN electron-positron collider, LEP, has been employed for the study of cosmic ray muons. The muon spectrometer of L3 consists of a set of high-precision drift chambers installed inside a magnet with a volume of about 1000 m**3 and a field of 0.5 T. Muon momenta are measured with a resolution of a few percent at 50 GeV. The detector is located under 30 m of overburden. A scintillator air shower array of 54 m by 30 m is installed on the roof of the surface hall above L3 in order to estimate the energy and the core position of the shower associated with a sample of detected muons. Thanks to the unique properties of the L3+C detector, muon research topics relevant to various current problems in cosmic ray and particle astrophysics can be studied.

  13. OT2_tvelusam_4: Probing Galactic Spiral Arm Tangencies with [CII

    Science.gov (United States)

    Velusamy, T.

    2011-09-01

    We propose to use the unique viewing geometry of the Galactic spiral arm tangents , which provide an ideal environment for studying the effects of density waves on spiral structure. We propose a well-sampled map of the[C II] 1.9 THz line emission along a 15-degree longitude region across the Norma-3kpc arm tangential, which includes the edge of the Perseus Arm. The COBE-FIRAS instrument observed the strongest [C II] and [N II] emission along these spiral arm tangencies.. The Herschel Open Time Key Project Galactic Observations of Terahertz C+ (GOT C+), also detects the strongest [CII] emission near these spiral arm tangential directions in its sparsely sampled HIFI survey of [CII] in the Galactic plane survey. The [C II] 158-micron line is the strongest infrared line emitted by the ISM and is an excellent tracer and probe of both the diffuse gases in the cold neutral medium (CNM) and the warm ionized medium (WIM). Furthermore, as demonstrated in the GOTC+ results, [C II] is an efficient tracer of the dark H2 gas in the ISM that is not traced by CO or HI observations. Thus, taking advantage of the long path lengths through the spiral arm across the tangencies, we can use the [C II] emission to trace and characterize the diffuse atomic and ionized gas as well as the diffuse H2 molecular gas in cloud transitions from HI to H2 and C+ to C and CO, throughout the ISM. The main goal of our proposal is to use the well sampled (at arcmin scale) [C II] to study these gas components of the ISM in the spiral-arm, and inter-arm regions, to constrain models of the spiral structure and to understand the influence of spiral density waves on the Galactic gas and the dynamical interaction between the different components. The proposed HIFI observations will consist of OTF 15 degree longitude scans and one 2-degree latitude scan sampled every 40arcsec across the Norma- 3kpc Perseus Spiral tangency.

  14. VHMPID RICH prototype using pressurized C{sub 4}F{sub 8}O radiator gas and VUV photon detector

    Energy Technology Data Exchange (ETDEWEB)

    Acconcia, T.V. [UNICAMP, University of Campinas, Campinas (Brazil); Agócs, A.G. [Wigner RCP of the HAS, Budapest (Hungary); Barile, F. [INFN Sezione di Bari and Universitá degli Studi di Bari, Dipartimento Interateneo di Fisica M. Merlin, Bari (Italy); Barnaföldi, G.G. [Wigner RCP of the HAS, Budapest (Hungary); Bellwied, R. [University of Houston, Houston (United States); Bencédi, G. [Wigner RCP of the HAS, Budapest (Hungary); Bencze, G., E-mail: Gyorgy.Bencze@cern.ch [Wigner RCP of the HAS, Budapest (Hungary); Berényi, D.; Boldizsár, L. [Wigner RCP of the HAS, Budapest (Hungary); Chattopadhyay, S. [Saha Institute, Kolkata (India); Chinellato, D.D. [University of Houston, Houston (United States); Cindolo, F. [University of Salerno, Salerno (Italy); Cossyleon, K. [Chicago State University, Chicago, IL (United States); Das, D.; Das, K.; Das-Bose, L. [Saha Institute, Kolkata (India); Dash, A.K. [UNICAMP, University of Campinas, Campinas (Brazil); D' Ambrosio, S. [University of Salerno, Salerno (Italy); De Cataldo, G. [INFN Sezione di Bari and Universitá degli Studi di Bari, Dipartimento Interateneo di Fisica M. Merlin, Bari (Italy); De Pasquale, S. [University of Salerno, Salerno (Italy); and others

    2014-12-11

    A small-size prototype of a new Ring Imaging Cherenkov (RICH) detector using for the first time pressurized C4F8O radiator gas and a photon detector consisting of MWPC equipped with a CsI photocathode has been built and tested at the PS accelerator at CERN. It contained all the functional elements of the detector proposed as Very High Momentum Particle Identification (VHMPID) upgrade for the ALICE experiment at LHC to provide charged hadron track-by-track identification in the momentum range starting from 5 potentially up to 25 GeV/c. In the paper the equipment and its elements are described and some characteristic test results are shown.

  15. Three-reflections telescope proposal as flat-field anastigmat for wide field observations at Dome C

    Science.gov (United States)

    Ferrari, M.; Lemaître, G.; Viotti, R.; La Padula, C.; Comte, G.; Blanc, M.; Boer, M.

    It is now evident that the exceptional seeing at Dome C will allow, in the next years, to pursue astronomical programs with conditions better than at any other observatory in the world, and very close to space experiments. Considering a new type of wide-field telescope, particular astronomical programs could be well optimized for observations at Dome C such as surveys for the discovery and follow up of near-Earth asteroids, search for extra-solar planets using transit or micro-lensing events, and stellar luminosity variations. We propose to build a 1.5 2m class three-reflections telescope, with 1 1.5degree FOV, four times shorter than an equivalent Schmidt telescope, and providing a flat field without requiring a triplet- or quadruplet-lens corrector since its design is anastigmatic. We present the preliminary optical tests of such designs: MINITRUST1 and 2 are two 45cm identical prototypes based in France and Italy, and manufactured using active optics techniques.

  16. High-Resolution C-Arm CT and Metal Artifact Reduction Software: A Novel Imaging Modality for Analyzing Aneurysms Treated with Stent-Assisted Coil Embolization.

    Science.gov (United States)

    Yuki, I; Kambayashi, Y; Ikemura, A; Abe, Y; Kan, I; Mohamed, A; Dahmani, C; Suzuki, T; Ishibashi, T; Takao, H; Urashima, M; Murayama, Y

    2016-02-01

    Combination of high-resolution C-arm CT and novel metal artifact reduction software may contribute to the assessment of aneurysms treated with stent-assisted coil embolization. This study aimed to evaluate the efficacy of a novel Metal Artifact Reduction prototype software combined with the currently available high spatial-resolution C-arm CT prototype implementation by using an experimental aneurysm model treated with stent-assisted coil embolization. Eight experimental aneurysms were created in 6 swine. Coil embolization of each aneurysm was performed by using a stent-assisted technique. High-resolution C-arm CT with intra-arterial contrast injection was performed immediately after the treatment. The obtained images were processed with Metal Artifact Reduction. Five neurointerventional specialists reviewed the image quality before and after Metal Artifact Reduction. Observational and quantitative analyses (via image analysis software) were performed. Every aneurysm was successfully created and treated with stent-assisted coil embolization. Before Metal Artifact Reduction, coil loops protruding through the stent lumen were not visualized due to the prominent metal artifacts produced by the coils. These became visible after Metal Artifact Reduction processing. Contrast filling in the residual aneurysm was also visualized after Metal Artifact Reduction in every aneurysm. Both the observational (P software. The combination of high-resolution C-arm CT and Metal Artifact Reduction enables differentiation of the coil mass, stent, and contrast material on the same image by significantly reducing the metal artifacts produced by the platinum coils. This novel image technique may improve the assessment of aneurysms treated with stent-assisted coil embolization. © 2016 by American Journal of Neuroradiology.

  17. The Aerogel Čerenkov detector for the SHMS magnetic spectrometer in Hall C at Jefferson Lab

    Science.gov (United States)

    Horn, T.; Mkrtchyan, H.; Ali, S.; Asaturyan, A.; Carmignotto, M.; Dittmann, A.; Dutta, D.; Ent, R.; Hlavin, N.; Illieva, Y.; Mkrtchyan, A.; Nadel-Turonski, P.; Pegg, I.; Ramos, A.; Reinhold, J.; Sapkota, I.; Tadevosyan, V.; Zhamkochyan, S.; Wood, S. A.

    2017-01-01

    Hadronic reactions producing strange quarks such as the exclusive p (e , e ‧K+) Λ and p (e , e ‧K+)Σ0 reactions, or the semi-inclusive p (e , e ‧K+) X reaction, play an important role in studies of hadron structure and the dynamics that bind the most basic elements of nuclear physics. The small-angle capability of the new Super High Momentum Spectrometer (SHMS) in Hall C, coupled with its high momentum reach - up to the anticipated 11-GeV beam energy in Hall C - and coincidence capability with the well-understood High Momentum Spectrometer (HMS), will allow for probes of such hadron structure involving strangeness down to the smallest distance scales to date. To cleanly select the kaons, a threshold aerogel Cerenkov detector has been constructed for the SHMS. The detector consists of an aerogel tray followed by a diffusion box. Four trays for aerogel of nominal refractive indices of n=1.030, 1.020, 1.015 and 1.011 were constructed. The tray combination will allow for identification of kaons from 1 GeV/c up to 7.2 GeV/c, reaching ∼10-2 proton and 10-3 pion rejection, with kaon detection efficiency better than 95%. The diffusion box of the detector is equipped with 14 five-inch diameter photomultiplier tubes. Its interior walls are covered with Gore diffusive reflector, which is superior to the commonly used Millipore paper and improved the detector performance by 35%. The inner surface of the two aerogel trays with higher refractive index is covered with Millipore paper, however, those two trays with lower aerogel refractive index are again covered with Gore diffusive reflector for higher performance. The measured mean number of photoelectrons in saturation is ∼12 for n=1.030, ∼8 for n=1.020, ∼10 for n=1.015, and ∼5.5 for n=1.011. The design details, the results of component characterization, and initial performance tests and optimization of the detector are presented.

  18. Evaluation of patient exposure with Flat Panel Detector (FPD) in X-ray TV system

    International Nuclear Information System (INIS)

    Yamada, M.; Komiya, N.; Kawaguchi, A.; Suzuki, M.; Suzuki, Shoichi; Asada, Yasuki

    2008-01-01

    The use of flat-panel detector (FPD) systems in TV equipment for gastrointestinal tract examination is increasing. The use of FPD systems is believed to reduce the exposure dose. When our institution changed its TV equipment from an image intensifier (GE; MS90Tj) system to an FPD (Shimadzu; SONIALVISION safire DAR-3500) system, we measured the doses produced and carried out a comparative examination of the extent to which exposure could be reduced. Two TV systems were used. We used an analyzer to measure output waveform, tube voltage, and half-value layer (HVL), and an ionization chamber dosimeter to carry out dose-in-air measurements. Body thickness, number of image acquisitions, and fluoroscopy time are required for the calculation of entrance skin dose (ESD). We therefore measured body thicknesses in 1000 upper gastrointestinal tract (UGI) and barium enemas and obtained average body thicknesses for males and females by age group. Values used for number of image acquisitions and fluoroscopy times were the averages in our institution over a two-year period. When an I.I. system was used, the average ESD during UGI examination were 126.8 mGy fluoroscopy dose and 11.62 mGy imaging dose, for an average total dose of 138.42 mGy per examination. ESD during barium enema averaged 201.73 mGy fluoroscopy dose and 45.2 mGy imaging dose, for an average total dose of 246.92 mGy per examination. When an FPD system was used, the average ESD during UGI examination were 58.71 mGy fluoroscopy dose and 5.72 mGy imaging dose, for an average total dose of 64.43 mGy per examination. ESD during barium enema averaged 112.21 mGy fluoroscopy dose and 24.55 mGy imaging dose, for an average total dose of 136.76 mGy per examination. The use of an FPD system reduced both fluoroscopy dose and imaging dose by 50%. The number of TV systems equipped with FPD in Japan has increased from around 1300 in 2006 to around 1700 in 2007. The use of FPD systems can be expected to increase in future. This

  19. Readout and Trigger for the AFP Detector at the ATLAS Experiment

    CERN Document Server

    Kocian, Martin; The ATLAS collaboration

    2018-01-01

    AFP, the ATLAS Forward Proton consists of silicon detectors at 205 m and 217 m on each side of ATLAS. In 2016 two detectors in one side were installed. The FEI4 chips are read at 160 Mbps over the optical fibers. The DAQ system uses a FPGA board with Artix chip and a mezzanine card with RCE data processing module based on a Zynq chip with ARM processor running Linux. In this contribution we give an overview of the AFP detector with the commissioning steps taken to integrate with the ATLAS TDAQ. Furthermore first performance results are presented.

  20. TH-C-17A-06: A Hardware Implementation and Evaluation of Robotic SPECT: Toward Molecular Imaging Onboard Radiation Therapy Machines

    International Nuclear Information System (INIS)

    Yan, S; Touch, M; Bowsher, J; Yin, F; Cheng, L

    2014-01-01

    Purpose: To construct a robotic SPECT system and demonstrate its capability to image a thorax phantom on a radiation therapy flat-top couch. The system has potential for on-board functional and molecular imaging in radiation therapy. Methods: A robotic SPECT imaging system was developed utilizing a Digirad 2020tc detector and a KUKA KR150-L110 robot. An imaging study was performed with the PET CT Phantom, which includes 5 spheres: 10, 13, 17, 22 and 28 mm in diameter. Sphere-tobackground concentration ratio was 6:1 of Tc99m. The phantom was placed on a flat-top couch. SPECT projections were acquired with a parallel-hole collimator and a single pinhole collimator. The robotic system navigated the detector tracing the flat-top table to maintain the closest possible proximity to the phantom. For image reconstruction, detector trajectories were described by six parameters: radius-of-rotation, x and z detector shifts, and detector rotation θ, tilt ϕ and twist γ. These six parameters were obtained from the robotic system by calibrating the robot base and tool coordinates. Results: The robotic SPECT system was able to maneuver parallel-hole and pinhole collimated SPECT detectors in close proximity to the phantom, minimizing impact of the flat-top couch on detector-to-COR (center-ofrotation) distance. In acquisitions with background at 1/6th sphere activity concentration, photopeak contamination was heavy, yet the 17, 22, and 28 mm diameter spheres were readily observed with the parallel hole imaging, and the single, targeted sphere (28 mm diameter) was readily observed in the pinhole region-of-interest (ROI) imaging. Conclusion: Onboard SPECT could be achieved by a robot maneuvering a SPECT detector about patients in position for radiation therapy on a flat-top couch. The robot inherent coordinate frame could be an effective means to estimate detector pose for use in SPECT image reconstruction. PHS/NIH/NCI grant R21-CA156390-01A1

  1. Rapid learning of magnetic compass direction by C57BL/6 mice in a 4-armed 'plus' water maze.

    Directory of Open Access Journals (Sweden)

    John B Phillips

    Full Text Available Magnetoreception has been demonstrated in all five vertebrate classes. In rodents, nest building experiments have shown the use of magnetic cues by two families of molerats, Siberian hamsters and C57BL/6 mice. However, assays widely used to study rodent spatial cognition (e.g. water maze, radial arm maze have failed to provide evidence for the use of magnetic cues. Here we show that C57BL/6 mice can learn the magnetic direction of a submerged platform in a 4-armed (plus water maze. Naïve mice were given two brief training trials. In each trial, a mouse was confined to one arm of the maze with the submerged platform at the outer end in a predetermined alignment relative to magnetic north. Between trials, the training arm and magnetic field were rotated by 180(° so that the mouse had to swim in the same magnetic direction to reach the submerged platform. The directional preference of each mouse was tested once in one of four magnetic field alignments by releasing it at the center of the maze with access to all four arms. Equal numbers of responses were obtained from mice tested in the four symmetrical magnetic field alignments. Findings show that two training trials are sufficient for mice to learn the magnetic direction of the submerged platform in a plus water maze. The success of these experiments may be explained by: (1 absence of alternative directional cues (2, rotation of magnetic field alignment, and (3 electromagnetic shielding to minimize radio frequency interference that has been shown to interfere with magnetic compass orientation of birds. These findings confirm that mice have a well-developed magnetic compass, and give further impetus to the question of whether epigeic rodents (e.g., mice and rats have a photoreceptor-based magnetic compass similar to that found in amphibians and migratory birds.

  2. Development of ultracold neutron detectors and a polarization analyzing system for the measurement of the neutron electric dipole moment

    International Nuclear Information System (INIS)

    Rogel, Gwendal

    2009-01-01

    This thesis was performed in the context of a project aiming to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute. Two aspects have been studied: The detection and the polarization analysis of ultracold neutrons. Three types of detectors have been tested at the Institut Laue-Langevin (ILL): The Cascade-U (GEM technology), the "3He gas detector and "6Li-doped glass scintillators (GS family). Their detection efficiency and their background sensitivity have been measured. The GS10 scintillator is competitive with the "3He gas detector under the conditions realized with the EDM spectrometer. A GS3/GS20 scintillator stack has enabled to improve the neutron/gamma discrimination. It has been found 20% less efficient than the "3He gas detector under the EDM spectrometer. The Cascade-U detector has been observed to be 20% less efficient than a 500 microns thick GS10 glass as confirmed by simulations. A new system for simultaneous spin analysis is presented. It consists of two independent detection systems (arms) which are each made of an adiabatic spin flipper, a spin analyzer, and a detector. The arms detect opposite spin components, allowing the simultaneous counting of both neutron spin orientations. A prototype mounted in horizontal configuration has been tested at ILL. The analyzing power of both arms has been measured to be 80%. The transmission of the system without spin analyzers has been found to be 50%. (author) [fr

  3. Flat detector CT and its applications in the endovascular treatment of wide-necked intracranial aneurysms—A literature review

    Energy Technology Data Exchange (ETDEWEB)

    Zwarzany, Łukasz, E-mail: zwarzany@gmail.com [Department of Radiology and Diagnostic Imaging, M. Curie Hospital, Arkońska 4, 71-455 Szczecin (Poland); Poncyljusz, Wojciech, E-mail: wponcyl@poczta.onet.pl [Departament of Interventional Radiology, Pomeranian Medical University, Neurointerventional Cath Lab MSW Hospital, Al. Powst. Wielkopolskich 72, 70-111 Szczecin (Poland); Burke, Thomas H., E-mail: tom.burke@microvention.com [Microvention – Terumo, Tustin, CA (United States)

    2017-03-15

    Highlights: • FDCT provides cross-sectional images of the intraluminal devices. • Contrast agent administration enables the evaluation of the surrounding vessels. • Artefacts generated by the coil mass are the major limitation of this imaging tool. • MAR software improves significantly the quality of the acquired images. • The examination can be performed during the procedure as well as in the follow-up. - Abstract: Flat detector CT (FDCT) provides cross sectional imaging within an angiographic suite and is increasingly gaining popularity in various areas of interventional radiology, as an alternative imaging modality. Its relatively high spatial resolution improves visualization of intraluminal devices such as intracranial stents or flow-diverters. Device deployment and positioning, in relation to the parent vessel and surrounding structures, are easily assessible with FDCT. Furthermore, with contrast agent administration, it expands the diagnostic capabilities of this new imaging tool. However, beam-hardening artifacts is a major limitation in some cases. The examination can be performed both during the endovascular procedure and for pre- and post-treatment imaging. Intravenous contrast agent injection reduces the risk of complications, making it possible to perform this examination in the outpatient settings. The aim of this paper is to present an overview of published studies reporting experience with FDCT in the field of endovascular neurosurgery and in particular, FDCT's contribution in treatment of wide-necked intracranial aneurysms. The authors have focused specifically on stent-assisted coiling and flow-diverter implantation, since obtaining proper parent vessel wall apposition of these devices is essential for short- and long-term procedural outcomes.

  4. Post procedural complications of cardiac implants done in a resource limited setting under 'C' arm: A single centre experience.

    Science.gov (United States)

    Jayachandra, A; Aggarwal, Vivek; Kumar, Sandeep; Nagesh, I V

    2018-04-01

    Cardiology interventions in peripheral hospitals is a challenging task where cardiologist have to fight against time and limited resources. Most of the sudden cardiac deaths occur due to arrhythmia and heart blocks/sinus node dysfunction. Our study is a single peripheral center experience of cardiac devices implantation using a 'C' Arm. The aim of this study was to post procedural complications of cardiac implants done in aresource limited setting under 'C' arm. This study is done at a peripheral cardiology center with no cardiac catheterization laboratory (CCL) facilities. Consecutive patients reporting to cardiology center, between Jan 2015 and Oct 2016, with a definite indication for cardiac device implant were included in the study. All the procedure of implantation was done in the operation theatre under 'C' arm under local anesthesia with continuous cardiac monitoring and critical care back up. Total 58 device implantations were done from Jan 2015 to Oct 2016. The mean age of the patients was 67.15 ± 10.85 years. Males constituted almost two third (68.9%) of patients. The commonest indication for device implantation was sinus node dysfunction in 60.34% followed by complete heart block in 25.86% and ventricular tachycardia in 12.06%. No post procedure infection was observed in our study. Device implantation constitute a major group of life saving interventions in cardiology practice. Our study has emphasised that when appropriate aseptic measures are taken during device implantation at peripheral centres, the complications rate are comparable to interventions done at advance cardiac centres.

  5. Radiation detector

    International Nuclear Information System (INIS)

    Conrad, B.; Finkenzeller, J.; Kiiehn, G.; Lichtenberg, W.

    1984-01-01

    In an exemplary embodiment, a flat radiation beam is detected having a common electrode disposed parallel to the beam plane at one side and a common support with a series of individual conductors providing electrodes opposite successive portions of the common electrode and lying in a plane also parallel to the beam plane. The beam may be fan-shaped and the individual electrodes may be aligned with respective ray paths separated by uniform angular increments in the beam plane. The individual conductors and the connection thereof to the exterior of the detector housing may be formed on an insulator which can be folded into a T-shape for leading the supply conductors for alternate individual conductors toward terminals at opposite sides of the chamber

  6. A Measurement of the B ---> Eta/C K Branching Fraction Using the BaBar Detector

    Energy Technology Data Exchange (ETDEWEB)

    Jackson, Frank; /Manchester U.

    2006-04-26

    The branching fraction is measured for the decay channels B{sup 0} {yields} {eta}{sub c}K{sub S}{sup 0} and B{sup +} {yields} {eta}{sub c}K{sup +} where {eta}{sub c} {yields} K{bar K}{pi}, using the BABAR detector. The {eta}{sub c} {yields} K{sub S}{sup 0}K{sup +}{pi}{sup -} and {eta}{sub c} {yields} K{sup +}K{sup -}{pi}{sup 0} decay channels are used, including non-resonant decays and possibly those through intermediate resonances.

  7. Uncooled Radiation Hard SiC Schottky VUV Detectors Capable of Single Photon Sensing, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — This project seeks to design, fabricate, characterize and commercialize very large area, uncooled and radiative hard 4H-SiC VUV detectors capable of near single...

  8. Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT

    International Nuclear Information System (INIS)

    Doelken, M.; Struffert, T.; Richter, G.; Engelhorn, T.; Doerfler, A.; Nimsky, C.; Ganslandt, O.; Hammen, T.

    2008-01-01

    The aim of this study was to compare flat-panel volumetric CT (VCT) to conventional CT (cCT) in the visualization of the extent of subarachnoid hemorrhage (SAH) and the width of the ventricles in patients with acute SAH. Included in the study were 22 patients with an acutely ruptured cerebral aneurysm who received VCT during coil embolization. VCT image quality, the extent of SAH (using a modified Fisher score and total slice number with SAH visible) and the width of the ventricles (Evans index) were evaluated by two experienced neuroradiologists (RAD1 and RAD2) and compared to the findings on cCT. Ten patients undergoing VCT for reasons other than SAH served as negative controls. Interobserver agreement in rating image quality was excellent for cCT (Kendall W value 0.94) and good for VCT (0.74). SAH was identified by RAD1 and RAD2 on VCT images in all patients. The modified Fisher scores underestimated the extent of SAH on VCT images in comparison with cCT images. Pearson's correlation coefficient (r) regarding the number of image slices with SAH visible on cCT images compared with the number on VCT images was 0.85 for RAD1 and 0.84 for RAD2. The r value for the degree of interobserver agreement for the number of slices with SAH visible was 0.99 for cCT, and 0.95 for VCT images (n 19), respectively. The width of the ventricles measured in terms of the Evans Index showed excellent concordance between the modalities (r = 0.81 vs. 0.82). Our preliminary results indicate that VCT is helpful in evaluating SAH in the angiography suite. Additionally, reliable evaluation of ventricle width is feasible. However, there are limitations with regard to the visibility of SAH on VCT images in comparison to cCT images. (orig.)

  9. Gas detector with a μm size strips anode

    International Nuclear Information System (INIS)

    Oed, A.

    1988-01-01

    A flat electrode device for an ionizing radiation multidetector, particularly for an X-ray detector used in tomodensitometry, is presented. It consists of either two active electrodes of the same kind, or an anode-electrode and a cathode electrode, on opposite sides of a base plate. The device avoids problems linked to flatness and parallelism, and the base plate consists of at least two intermediate plates separated by a space containing at least layer of binding material. The device thus overcomes difficulties associated with thickness and the need to stop ionizing radiation from passing from one cell to another by traversing the base plate. The steps of the fabrication process are detailed [fr

  10. Irradiation of 4H-SiC UV detectors with heavy ions

    International Nuclear Information System (INIS)

    Kalinina, E. V.; Lebedev, A. A.; Bogdanova, E.; Berenquier, B.; Ottaviani, L.; Violina, G. N.; Skuratov, V. A.

    2015-01-01

    Ultraviolet (UV) photodetectors based on Schottky barriers to 4H-SiC are formed on lightly doped n-type epitaxial layers grown by the chemical vapor deposition method on commercial substrates. The diode structures are irradiated at 25°C by 167-MeV Xe ions with a mass of 131 amu at a fluence of 6 × 10 9 cm −2 . Comparative studies of the optical and electrical properties of as-grown and irradiated structures with Schottky barriers are carried out in the temperature range 23–180°C. The specific features of changes in the photosensitivity and electrical characteristics of the detector structures are accounted for by the capture of photogenerated carriers into traps formed due to fluctuations of the conduction-band bottom and valence-band top, with subsequent thermal dissociation

  11. Fully solution processed Al-TiO2-Si (MIS) structured photo-detector

    Science.gov (United States)

    Mondal, Sandip; Kumar, Arvind

    2018-05-01

    We demonstrate the fabrication of a high performance photo detector by fully solution processed technique. The detector is fabricated with photo sensitive, low temperature (200˚C) and sol-gel processed titanium dioxide (TiO2) dielectric material on silicon substrate in the form of MIS structure with top aluminum gate. The optical detection experiment is performed on Al—TiO2—Si (MIS) device by measuring the capacitance—voltage (CV at 100 kHz) curve within the visible region of light (365 — 700 nm). The presence of light shift the flat band voltage (VFB) from 290 mV to 360 mV due to the generation of photo activated charge carriers by UV (365 nm) and white light, respectively. Moreover, the generation of the charge carrier increases drastically by the combination of UV and white, which resulting as a very large shift (600 mV) in the VFB. The entire experiment was performed in normal lab conditions with open air environment, without any clean room facility.

  12. High resolution micro-CT of low attenuating organic materials using large area photon-counting detector

    International Nuclear Information System (INIS)

    Kumpová, I.; Jandejsek, I.; Jakůbek, J.; Vopálenský, M.; Vavřík, D.; Fíla, T.; Koudelka, P.; Kytýř, D.; Zlámal, P.; Gantar, A.

    2016-01-01

    To overcome certain limitations of contemporary materials used for bone tissue engineering, such as inflammatory response after implantation, a whole new class of materials based on polysaccharide compounds is being developed. Here, nanoparticulate bioactive glass reinforced gelan-gum (GG-BAG) has recently been proposed for the production of bone scaffolds. This material offers promising biocompatibility properties, including bioactivity and biodegradability, with the possibility of producing scaffolds with directly controlled microgeometry. However, to utilize such a scaffold with application-optimized properties, large sets of complex numerical simulations using the real microgeometry of the material have to be carried out during the development process. Because the GG-BAG is a material with intrinsically very low attenuation to X-rays, its radiographical imaging, including tomographical scanning and reconstructions, with resolution required by numerical simulations might be a very challenging task. In this paper, we present a study on X-ray imaging of GG-BAG samples. High-resolution volumetric images of investigated specimens were generated on the basis of micro-CT measurements using a large area flat-panel detector and a large area photon-counting detector. The photon-counting detector was composed of a 010× 1 matrix of Timepix edgeless silicon pixelated detectors with tiling based on overlaying rows (i.e. assembled so that no gap is present between individual rows of detectors). We compare the results from both detectors with the scanning electron microscopy on selected slices in transversal plane. It has been shown that the photon counting detector can provide approx. 3× better resolution of the details in low-attenuating materials than the integrating flat panel detectors. We demonstrate that employment of a large area photon counting detector is a good choice for imaging of low attenuating materials with the resolution sufficient for numerical

  13. High resolution micro-CT of low attenuating organic materials using large area photon-counting detector

    Science.gov (United States)

    Kumpová, I.; Vavřík, D.; Fíla, T.; Koudelka, P.; Jandejsek, I.; Jakůbek, J.; Kytýř, D.; Zlámal, P.; Vopálenský, M.; Gantar, A.

    2016-02-01

    To overcome certain limitations of contemporary materials used for bone tissue engineering, such as inflammatory response after implantation, a whole new class of materials based on polysaccharide compounds is being developed. Here, nanoparticulate bioactive glass reinforced gelan-gum (GG-BAG) has recently been proposed for the production of bone scaffolds. This material offers promising biocompatibility properties, including bioactivity and biodegradability, with the possibility of producing scaffolds with directly controlled microgeometry. However, to utilize such a scaffold with application-optimized properties, large sets of complex numerical simulations using the real microgeometry of the material have to be carried out during the development process. Because the GG-BAG is a material with intrinsically very low attenuation to X-rays, its radiographical imaging, including tomographical scanning and reconstructions, with resolution required by numerical simulations might be a very challenging task. In this paper, we present a study on X-ray imaging of GG-BAG samples. High-resolution volumetric images of investigated specimens were generated on the basis of micro-CT measurements using a large area flat-panel detector and a large area photon-counting detector. The photon-counting detector was composed of a 010× 1 matrix of Timepix edgeless silicon pixelated detectors with tiling based on overlaying rows (i.e. assembled so that no gap is present between individual rows of detectors). We compare the results from both detectors with the scanning electron microscopy on selected slices in transversal plane. It has been shown that the photon counting detector can provide approx. 3× better resolution of the details in low-attenuating materials than the integrating flat panel detectors. We demonstrate that employment of a large area photon counting detector is a good choice for imaging of low attenuating materials with the resolution sufficient for numerical simulations.

  14. A practical method for three-dimensional reconstruction of joints using a C-arm system and shift-and-add algorithm

    International Nuclear Information System (INIS)

    Li Senhu; Jiang Huabei

    2005-01-01

    Currently, radiography with C-arm systems is playing a major role in the assessment of arthritis. However, the radiographic two-dimensional projection images of joints often interfere with physicians' efforts to better understand and measure the structure changes of joints due to the overlap of bone structures at different depths. An accurate, low-cost, and practical three-dimensional (3D) reconstruction approach of joints will be beneficial in diagnosing arthritis. Toward this end, a novel method is developed in this paper based on a C-arm system. The idea is to apply the shift-and-add algorithm (commonly used in digital tomosynthesis) on the segmented projection images at multiple angles, which results in accurate reconstruction of the 3D structures of joints. The method provides a new solution to precisely distinguish objects from blurring background. The proposed method has been tested and evaluated on simulated cylinders, a chicken bone phantom with known structure, and an in vivo human index finger. The results are demonstrated and discussed

  15. Characterisation and commissioning of the LHCb VELO detector

    CERN Document Server

    AUTHOR|(CDS)2069345; Jans, E

    2009-01-01

    The LHCb detector is a one-armed spectrometer at the Large Hadron Collider. It has been designed to look for physics beyond the Standard Model through high precision measurements of CP-violation in the B-system and through the detection of rare B-decays. The success of LHCb relies heavily on its vertex detector, the VELO (VErtex LOcator), which will be used to trigger on B decay vertices and reconstruct them with micrometre accuracy. While small in size, comprising just 84 sensors, the VELO construction poses special challenges due to the high accuracy required and the proximity to the LHC beams. The detector will be required to operate under vacuum and will be exposed to high radiation levels. The first section of the thesis gives a brief introduction to the LHCb detector and the physics programme of the LHCb collaboration. The following sections review the VELO design and give and in-depth report on measurements of the detector performance based on data collected in beam tests. Topics covered are hit resolu...

  16. The aerogel threshold Cherenkov detector for the high momentum spectrometer in Hall C at Jefferson lab

    International Nuclear Information System (INIS)

    Razmik Asaturyan; Rolf Ent; Howard Fenker; David Gaskell; Garth Huber; Mark Jones; David Mack; Hamlet Mkrtchyan; Bert Metzger; Nadia Novikoff; Vardan Tadevosyan; William Vulcan; Stephen Wood

    2004-01-01

    We describe a new aerogel threshold Cherenkov detector installed in the HMS spectrometer in Hall C at Jefferson Lab. The Hall C experimental program in 2003 required an improved particle identification system for better identification of π/K/p, which was achieved by installing an additional threshold Cherenkov counter. Two types of aerogel with n = 1.03 and n = 1.015 allow one to reach ∼10 -3 proton and 10 -2 kaon rejection in the 1-5 GeV/c momentum range with pion detection efficiency better than 99% (97%). The detector response shows no significant position dependence due to a diffuse light collection technique. The diffusion box was equipped with 16 Photonis XP4572 PMT's. The mean number of photoelectrons in saturation was ∼16 and ∼8, respectively. Moderate particle identification is feasible near threshold

  17. DC servo motor positioning with anti-windup implementation using C2000 ARM-Texas Instrument

    Science.gov (United States)

    Linggarjati, Jimmy

    2017-12-01

    One of the most important topics in control system is DC Motor. At this research, a positioning control system for a DC motor is investigated. Firstly, the DC Motor will be paramaterized to get the transfer function model, in order to be simulated in Matlab, and then implemented in a C2000-ARM microcontroller from TI (Texas Instrument). With this investigation, students in control system theory will be able to understand the importance of classical control theories, in relation to the real world implementation of the position control for the DC Motor, escpecially the importance of Anti-Windup technique in real-world implementation.

  18. Acquiring Multiview C-Arm Images to Assist Cardiac Ablation Procedures

    Directory of Open Access Journals (Sweden)

    Fallavollita Pascal

    2010-01-01

    Full Text Available CARTO XP is an electroanatomical cardiac mapping system that provides 3D color-coded maps of the electrical activity of the heart; however it is expensive and it can only use a single costly magnetic catheter for each patient intervention. Our approach consists of integrating fluoroscopic and electrical data from the RF catheters into the same image so as to better guide RF ablation, shorten the duration of this procedure, increase its efficacy, and decrease hospital cost when compared to CARTO XP. We propose a method that relies on multi-view C-arm fluoroscopy image acquisition for (1 the 3D reconstruction of the anatomical structure of interest, (2 the robust temporal tracking of the tip-electrode of a mapping catheter between the diastolic and systolic phases and (3 the 2D/3D registration of color coded isochronal maps directly on the 2D fluoroscopy image that would help the clinician guide the ablation procedure much more effectively. The method has been tested on canine experimental data.

  19. Second generation "Flat-Fish" vacuum chamber for an ISR intersection

    CERN Multimedia

    CERN PhotoLab

    1977-01-01

    The picture shows the part of the "Flat-Fish" chamber covering one side of the beam intersection region. The other side is covered by an equal part and both are welded toghether at their small end. The two shells of each part were hydroformed from 0.15 mm thick inconel 718 sheet (with end parts in inconel 600 for easier manual welding to the arms) and welded toghether with two strips which were attached by means of stainless steel strings to the Split Field Magnet poles in order to take the atmospheric pressure forces. This was the thinnest vacuum chamber made for the ISR. Inconel material was chosen for its high elastic modulus and strenght at chamber bake-out temperature.

  20. Current status and requirements for position-sensitive detectors in medicine

    CERN Document Server

    Speller, R

    2002-01-01

    This review considers the current status of detector developments for medical imaging using ionising radiation. This field is divided into two major areas; the use of X-rays for transmission imaging and the use of radioactive tracers in emission imaging (nuclear medicine). Until recently, most detector developments were for applications in nuclear medicine. However, in the past 5 years new developments in large area, X-ray-sensitive detectors have meant that both application domains are equally served. In X-ray imaging, work in CT and mammography are chosen as examples of sensor developments. Photodiode arrays in multi-slice spiral CT acquisitions are described and for mammography the use of amorphous silicon flat panel arrays is considered. The latter is an excellent example where new detector developments have required a re-think of traditional imaging methods. In gamma-ray imaging the recent developments in small area, task-specific cameras are described. Their limitations and current proposals to overcome...

  1. Detectors - Electronics

    International Nuclear Information System (INIS)

    Bregeault, J.; Gabriel, J.L.; Hierle, G.; Lebotlan, P.; Leconte, A.; Lelandais, J.; Mosrin, P.; Munsch, P.; Saur, H.; Tillier, J.

    1998-01-01

    The reports presents the main results obtained in the fields of radiation detectors and associated electronics. In the domain of X-ray gas detectors for the keV range efforts were undertaken to rise the detector efficiency. Multiple gap parallel plate chambers of different types as well as different types of X → e - converters were tested to improve the efficiency (values of 2.4% at 60 KeV were reached). In the field of scintillators a study of new crystals has been carried out (among which Lutetium orthosilicate). CdTe diode strips for obtaining X-ray imaging were studied. The complete study of a linear array of 8 CdTe pixels has been performed and certified. The results are encouraging and point to this method as a satisfying solution. Also, a large dimension programmable chamber was used to study the influence of temperature on the inorganic scintillators in an interval from -40 deg. C to +150 deg. C. Temperature effects on other detectors and electronic circuits were also investigated. In the report mentioned is also the work carried out for the realization of the DEMON neutron multidetector. For neutron halo experiments different large area Si detectors associated with solid and gas position detectors were realized. In the frame of a contract with COGEMA a systematic study of Li doped glasses was undertaken aiming at replacing with a neutron probe the 3 He counters presently utilized in pollution monitoring. An industrial prototype has been realised. Other studies were related to integrated analog chains, materials for Cherenkov detectors, scintillation probes for experiments on fundamental processes, gas position sensitive detectors, etc. In the field of associated electronics there are mentioned the works related to the multidetector INDRA, data acquisition, software gamma spectrometry, automatic gas pressure regulation in detectors, etc

  2. Creating flat design websites

    CERN Document Server

    Pratas, Antonio

    2014-01-01

    This book contains practical, step-by-step tutorials along with plenty of explanation about designing your flat website. Each section is introduced sequentially, building up your web design skills and completing your website.Creating Flat Design Websites is ideal for you if you are starting on your web development journey, but this book will also benefit seasoned developers wanting to start developing in flat.

  3. 40 CFR 230.42 - Mud flats.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 24 2010-07-01 2010-07-01 false Mud flats. 230.42 Section 230.42... Aquatic Sites § 230.42 Mud flats. (a) Mud flats are broad flat areas along the sea coast and in coastal rivers to the head of tidal influence and in inland lakes, ponds, and riverine systems. When mud flats...

  4. Flat-port connectors

    KAUST Repository

    Alrashed, Mohammed

    2017-05-26

    Disclosed are various embodiments for connectors used with electronic devices, such as input and/or output ports to connect peripheral equipment or accessories. More specifically, various flat-port are provided that can be used in place of standard connectors including, but not limited to, audio jacks and Universal Serial Bus (USB) ports. The flat-port connectors are an alternate connection design to replace the traditional receptacle port (female-port), making the device more sealed creation more dust and water resistant. It is unique in the way of using the outer surfaces of the device for the electrical connection between the ports. Flat-port design can allow the manufacture of extremely thin devices by eliminating the side ports slots that take a lot of space and contribute to the increase thickness of the device. The flat-port receptacle improves the overall appearance of the device and makes it more resistant to dust and water.

  5. Flat-space singletons

    International Nuclear Information System (INIS)

    Fronsdal, C.

    1987-01-01

    Singletons exist, as particles and as local fields, only in 3+2 de Sitter space. Their kinematical properties make them natural candidates for constituents of massless fields, and perhaps for quarks. It is interesting to find out how to describe this type of compositeness in flat space. A theory of interacting singleton fields in de Sitter space is now available, and in this paper we study the flat-space limit of the Green's functions of that theory. The flat-space limit is an autonomous theory of Green's functions, but is not an operator field theory. The three-point function is calculated and its flat-space limit is found to reveal glimpses of a physical interpretation. Causal and spectral properties are in accord with the tenets of axiomatic field theory. The theory is a generalization of local field theory, in which photons appear as composite objects although the physical S matrix is the same as in conventional QED

  6. Nano-optical observation of cascade switching in a parallel superconducting nanowire single photon detector

    International Nuclear Information System (INIS)

    Heath, Robert M.; Tanner, Michael G.; Casaburi, Alessandro; Hadfield, Robert H.; Webster, Mark G.; San Emeterio Alvarez, Lara; Jiang, Weitao; Barber, Zoe H.; Warburton, Richard J.

    2014-01-01

    The device physics of parallel-wire superconducting nanowire single photon detectors is based on a cascade process. Using nano-optical techniques and a parallel wire device with spatially separate pixels, we explicitly demonstrate the single- and multi-photon triggering regimes. We develop a model for describing efficiency of a detector operating in the arm-trigger regime. We investigate the timing response of the detector when illuminating a single pixel and two pixels. We see a change in the active area of the detector between the two regimes and find the two-pixel trigger regime to have a faster timing response than the one-pixel regime

  7. C-RED One and C-RED2: SWIR high-performance cameras using Saphira e-APD and Snake InGaAs detectors

    Science.gov (United States)

    Gach, Jean-Luc; Feautrier, Philippe; Stadler, Eric; Clop, Fabien; Lemarchand, Stephane; Carmignani, Thomas; Wanwanscappel, Yann; Boutolleau, David

    2018-02-01

    After the development of the OCAM2 EMCCD fast visible camera dedicated to advanced adaptive optics wavefront sensing, First Light Imaging moved to the SWIR fast cameras with the development of the C-RED One and the C-RED 2 cameras. First Light Imaging's C-RED One infrared camera is capable of capturing up to 3500 full frames per second with a subelectron readout noise and very low background. C-RED One is based on the last version of the SAPHIRA detector developed by Leonardo UK. This breakthrough has been made possible thanks to the use of an e-APD infrared focal plane array which is a real disruptive technology in imagery. C-RED One is an autonomous system with an integrated cooling system and a vacuum regeneration system. It operates its sensor with a wide variety of read out techniques and processes video on-board thanks to an FPGA. We will show its performances and expose its main features. In addition to this project, First Light Imaging developed an InGaAs 640x512 fast camera with unprecedented performances in terms of noise, dark and readout speed based on the SNAKE SWIR detector from Sofradir. The camera was called C-RED 2. The C-RED 2 characteristics and performances will be described. The C-RED One project has received funding from the European Union's Horizon 2020 research and innovation program under grant agreement N° 673944. The C-RED 2 development is supported by the "Investments for the future" program and the Provence Alpes Côte d'Azur Region, in the frame of the CPER.

  8. Performance quantification of a flat-panel imager in industrial mega-voltage X-ray imaging systems

    Energy Technology Data Exchange (ETDEWEB)

    Stritt, Carina, E-mail: carina.stritt@empa.ch [Empa, Swiss Federal Laboratories for Material Science and Technology, Center for X-ray Analytics, Ueberlandstrasse 129, 8600 Dübendorf (Switzerland); Plamondon, Mathieu; Hofmann, Jürgen; Flisch, Alexander [Empa, Swiss Federal Laboratories for Material Science and Technology, Center for X-ray Analytics, Ueberlandstrasse 129, 8600 Dübendorf (Switzerland); Sennhauser, Urs [Empa, Swiss Federal Laboratories for Material Science and Technology, Reliability Science and Technology Laboratory, Ueberlandstrasse 129, 8600 Dübendorf (Switzerland)

    2017-03-11

    Active matrix flat-panel detectors have gained popularity amongst X-ray imaging systems due to their speed, resolution and high dynamic range. With appropriate shielding modern flat-panel imagers can even be used in high energy Computed Tomography (CT) systems of energies up to several mega-electronvolt (MeV). However, the performance of a digital detector is not independent of the rest of the radiographic system but depends on all other components of the system. Signal and noise transfer properties highly depend on all parameters of an imaging chain. This work focuses on quantifying the resolution capabilities and the noise in the signals of a MeV X-ray imaging system. The performance quantification is done by computing the modulation transfer function (MTF) using the standard edge method as well as the noise power spectrum (NPS) of the imaging system. We performed Monte Carlo (MC) simulations in order to understand the influence of scattered radiation on the measurements. A comparison of the horizontal and vertical MTF showed that the imaging behaviour of the detector is isotropic. Moreover, an additional investigation of the noise performance of the system showed that there is no measurable noise correlation present in the system. It was shown that the thickness of the edge device does not have a significant influence on the resulting system MTF. A rapid drop in the visibility could be observed resulting in a value of 1.2 line pairs per mm at 50% MTF. The visibility limit of line pair patterns was found to be at 2.3 line pairs per mm given by the 10% MTF value.

  9. Performance quantification of a flat-panel imager in industrial mega-voltage X-ray imaging systems

    International Nuclear Information System (INIS)

    Stritt, Carina; Plamondon, Mathieu; Hofmann, Jürgen; Flisch, Alexander; Sennhauser, Urs

    2017-01-01

    Active matrix flat-panel detectors have gained popularity amongst X-ray imaging systems due to their speed, resolution and high dynamic range. With appropriate shielding modern flat-panel imagers can even be used in high energy Computed Tomography (CT) systems of energies up to several mega-electronvolt (MeV). However, the performance of a digital detector is not independent of the rest of the radiographic system but depends on all other components of the system. Signal and noise transfer properties highly depend on all parameters of an imaging chain. This work focuses on quantifying the resolution capabilities and the noise in the signals of a MeV X-ray imaging system. The performance quantification is done by computing the modulation transfer function (MTF) using the standard edge method as well as the noise power spectrum (NPS) of the imaging system. We performed Monte Carlo (MC) simulations in order to understand the influence of scattered radiation on the measurements. A comparison of the horizontal and vertical MTF showed that the imaging behaviour of the detector is isotropic. Moreover, an additional investigation of the noise performance of the system showed that there is no measurable noise correlation present in the system. It was shown that the thickness of the edge device does not have a significant influence on the resulting system MTF. A rapid drop in the visibility could be observed resulting in a value of 1.2 line pairs per mm at 50% MTF. The visibility limit of line pair patterns was found to be at 2.3 line pairs per mm given by the 10% MTF value.

  10. Proprioceptive Interaction between the Two Arms in a Single-Arm Pointing Task.

    Directory of Open Access Journals (Sweden)

    Kazuyoshi Kigawa

    Full Text Available Proprioceptive signals coming from both arms are used to determine the perceived position of one arm in a two-arm matching task. Here, we examined whether the perceived position of one arm is affected by proprioceptive signals from the other arm in a one-arm pointing task in which participants specified the perceived position of an unseen reference arm with an indicator paddle. Both arms were hidden from the participant's view throughout the study. In Experiment 1, with both arms placed in front of the body, the participants received 70-80 Hz vibration to the elbow flexors of the reference arm (= right arm to induce the illusion of elbow extension. This extension illusion was compared with that when the left arm elbow flexors were vibrated or not. The degree of the vibration-induced extension illusion of the right arm was reduced in the presence of left arm vibration. In Experiment 2, we found that this kinesthetic interaction between the two arms did not occur when the left arm was vibrated in an abducted position. In Experiment 3, the vibration-induced extension illusion of one arm was fully developed when this arm was placed at an abducted position, indicating that the brain receives increased proprioceptive input from a vibrated arm even if the arm was abducted. Our results suggest that proprioceptive interaction between the two arms occurs in a one-arm pointing task when the two arms are aligned with one another. The position sense of one arm measured using a pointer appears to include the influences of incoming information from the other arm when both arms were placed in front of the body and parallel to one another.

  11. Module production of the one-arm AFP 3D pixel tracker

    CERN Document Server

    Grinstein, S.; Chmeissani, M.; Dorholt, O.; Förster, F.; Lange, J.; Lopez Paz, I.; Manna, M.; Pellegrini, G.; Quirion, D.; Rijssenbeek, M.; Rohne, O.; Stugu, B.

    2016-01-01

    The ATLAS Forward Proton (AFP) detector is designed to identify events in which one or two protons emerge intact from the LHC collisions. AFP will consist of a tracking detector, to measure the momentum of the protons, and a time of flight system to reduce the background from multiple proton-proton interactions. Following an extensive qualification period, 3D silicon pixel sensors were selected for the AFP tracker. The sensors were produced at CNM (Barcelona) during 2014. The tracker module assembly and quality control was performed at IFAE during 2015. The assembly of the first AFP arm and the following installation in the LHC tunnel took place in February 2016. This paper reviews the fabrication process of the AFP tracker focusing on the pixel modules.

  12. On Recall Rate of Interest Point Detectors

    DEFF Research Database (Denmark)

    Aanæs, Henrik; Dahl, Anders Lindbjerg; Pedersen, Kim Steenstrup

    2010-01-01

    in relation to the number of interest points, the recall rate as a function of camera position and light variation, and the sensitivity relative to model parameter change. The overall conclusion is that the Harris corner detector has a very high recall rate, but is sensitive to change in scale. The Hessian......In this paper we provide a method for evaluating interest point detectors independently of image descriptors. This is possible because we have compiled a unique data set enabling us to determine if common interest points are found. The data contains 60 scenes of a wide range of object types......, and for each scene we have 119 precisely located camera positions obtained from a camera mounted on an industrial robot arm. The scene surfaces have been scanned using structured light, providing precise 3D ground truth. We have investigated a number of the most popular interest point detectors. This is done...

  13. Robust flat bands in RCo5 (R=rare earth) compounds

    OpenAIRE

    Ochi, Masayuki; Arita, Ryotaro; Matsumoto, Munehisa; Kino, Hiori; Miyake, Takashi

    2014-01-01

    The mechanism to realize the peculiar flat bands generally existing in RCo5 (R=rare earth) compounds is clarified by analyzing the first-principles band structures and the tight-binding model. These flat bands are constructed from the localized eigenstates, the existence of which is guaranteed by the partial cancelation between the intersite hopping amplitudes among the Co-3d states at the Kagome sites and those between the Kagome and honeycomb sites. Their relative positions to other bands c...

  14. Lithium germanium detectors reactivation

    International Nuclear Information System (INIS)

    Nicolai, J.A.; Marti, G.V.; Riso, J.M.; Gimenez, C.R.

    1981-01-01

    A convenient method to regenerate the characteristics of damaged Ge(li) detectors, that has been applied in the authors' laboratory, is described. The procedure consists in warming-up the crystal in its cryostat to temperatures between 10 deg C and 30 deg C above room temperature, in order to clean its surface. Subsequent cooling down to liquid nitrogen temperature, followed by one or more clean-up drifting processes, are applied to the crystals. This paper summarizes the results obtained with several detectors; this method was applied successfully to 15 detectors more. (author) [es

  15. C-arm fluoroscopy: a reliable modality for retrieval of foreign bodies in the maxillofacial region.

    Science.gov (United States)

    Pandyan, Deepak; Nandakumar, N; Qayyumi, Burhanuddin N; Kumar, Santosh

    2013-11-01

    The anatomic complexity of the maxillofacial region makes the retrieval of foreign bodies a daunting task for the maxillofacial Surgeon. Moreover the inability of 2-dimensional imaging to precisely locate foreign bodies makes it challenging. The anatomic proximity of critical structures and esthetic considerations limits the access and thus poses a greater challenge for the surgeon in cases of foreign body retrieval. Hereby we propose a simple technique and a case report to support, the retrieval of small (mobile C arm Fluoroscopy and a needle triangulation method to precisely locate a loosened miniplate screw in the mandibular angle region.

  16. Flat slices in Minkowski space

    Science.gov (United States)

    Murchadha, Niall Ó.; Xie, Naqing

    2015-03-01

    Minkowski space, flat spacetime, with a distance measure in natural units of d{{s}2}=-d{{t}2}+d{{x}2}+d{{y}2}+d{{z}2}, or equivalently, with spacetime metric diag(-1, +1, +1, +1), is recognized as a fundamental arena for physics. The Poincaré group, the set of all rigid spacetime rotations and translations, is the symmetry group of Minkowski space. The action of this group preserves the form of the spacetime metric. Each t = constant slice of each preferred coordinate system is flat. We show that there are also nontrivial non-singular representations of Minkowski space with complete flat slices. If the embedding of the flat slices decays appropriately at infinity, the only flat slices are the standard ones. However, if we remove the decay condition, we find non-trivial flat slices with non-vanishing extrinsic curvature. We write out explicitly the coordinate transformation to a frame with such slices.

  17. Fabrication of wire and flat strips with elevated recrystallization temperature of Mo monocrystals

    International Nuclear Information System (INIS)

    Mikhajlov, S.M.; Nesgovorov, V.V.; Kabakova, L.G.; Korzukhin, V.E.; Savitskij, E.M.; Burkhanov, G.S.; Ottenberg, E.V.

    1977-01-01

    A technique is developed for manufacturing wire and flat strip of elevated recrystallization point from single crystals fo molybdenum with micro-additions of zirconium and titanium by rotary hot forging with subsequent drawing under hydrodynamic friction conditions. Flat strip is manufactured next from a wire annealed at 1300-1400 deg C in hydrogen. Resultant wire and flat strip feature a high recrystallization point and a good shape stability. Tests on their ultimate strength on the range of temperatures between 20 and 1700 deg C have shown that the maximum temperature of the recrystallization onset is that of a wire from Mo single crystals of orientation [110], containing micro-additions of Zr and Ti, whereas loss of strength is at its highest in a wire from non-alloyed single-crystal molybdenum

  18. Ultra-high resolution C-Arm CT arthrography of the wrist: Radiation dose and image quality compared to conventional multidetector computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Werncke, Thomas, E-mail: Werncke.Thomas@mh-hannover.de [Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover (Germany); Sonnow, Lena; Meyer, Bernhard C. [Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover (Germany); Lüpke, Matthias [University of Veterinary Medicine Hannover, Institute for General Radiology and Medical Physics, Bischofsholer Damm 15, 30173 Hannover (Germany); Hinrichs, Jan; Wacker, Frank K.; Falck, Christian von [Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover (Germany)

    2017-04-15

    Objective: Objective of this phantom and cadaveric study was to compare the effective radiation dose (ED) and image quality (IQ) between C-arm computed tomography (CACT) using an ultra-high resolution 1 × 1 binning with a standard 16-slice CT (MDCT) arthrography of the wrist. Methods: ED was determined with thermoluminescence dosimetry using an anthropomorphic phantom and different patient positions. Imaging was conducted in 10 human cadaveric wrists after tri-compartmental injection of diluted iodinated contrast material and a wire phantom. IQ of MDCT was compared with CACT reconstructed with a soft (CACT1) and sharp (CACT2) kernel. High and low contrast resolution was determined. Three radiologists assessed IQ of wrist structures and occurrence of image artifacts using a 5-point Likert scale. Results: ED of MDCT was comparable to standard CACT (4.3 μSv/3.7 μSv). High contrast resolution was best for CACT2, decreased to CACT1 and MDCT. Low contrast resolution increased between CACT2 and MDCT (P < 0.001). IQ was best for CACT2 (1.3 ± 0.5), decreased to CACT1 (1.9 ± 0.6) and MDCT (3.5 ± 0.6). Non-compromising artifacts were only reported for CACT. Conclusions: The results of this phantom and cadaveric study indicate that ultra-high resolution C-Arm CT arthrography of the wrist bears the potential to outperform MDCT arthrography in terms of image quality and workflow at the cost of mildly increasing image artifacts while radiation dose to the patient is comparably low for both, MDCT and C-Arm CT.

  19. Flat detector computed tomography in diagnostic and interventional pediatric cardiology; Flachdetektor-Computertomografie in der diagnostischen und interventionellen Kinderkardiologie

    Energy Technology Data Exchange (ETDEWEB)

    Moesler, J.; Dittrich, S.; Gloeckler, M. [Erlangen-Nuernberg Univ., Erlangen (Germany). Pediatric Cardiology; Rompel, O. [Erlangen-Nuernberg Univ., Erlangen (Germany). Radiology

    2013-05-15

    Purpose: In this study the use of flat detector computed tomography (FD-CT) in the catheterization of patients with congenital heart disease was evaluated. Application reports were created for various issues based on the achieved image quality in diverse anatomical regions. Materials and Methods: FD-CT was applied in 176 cases during catheterization between January 2010 and April 2012. A five-point Likert scale ('essential' to 'misleading') was used to evaluate image quality. All cases were analyzed retrospectively and application reports for the visualization of the aorta, pulmonary arteries, pulmonary veins, semilunar valves, cavopulmonary connections and atrial baffles were generated. Contrast dye consumption and radiation dose were evaluated. Results: During the observation period FD-CT was applied in all 176 cases. The mean patient age was 7.0 years (0.01 - 42.53 years). The clinical value of FD-CT was rated superior to conventional angiography in 96.6 % of the cases and was never rated as 'misleading'. FD-CT was rated 'essential' in 3.4 % of all cases, 'very useful' in 77.3 % of all cases, 'useful' in 15.9 % of all cases and 'not useful' in 3.4 % of all cases. The mean dose-area product was 99 {mu}Gym{sup 2} (19.3 - 1276.6 {mu}Gym{sup 2}), and the used contrast dye was 1.76 ml/kg (0.9 - 5 ml/kg). Application reports for the visualization of different anatomical regions are demonstrated. Conclusion: FD-CT is a new and auxiliary procedure in diagnostic and interventional catheterization of patients with congenital heart disease. Particularly extracardiac structures can be displayed in three-dimensional high resolution and be used for diagnosis, surgical planning and 3 D navigation. (orig.)

  20. Test results of a new detector system for gamma ray isotopic measurements

    International Nuclear Information System (INIS)

    Malcom, J.E.; Bonner, C.A.; Hurd, J.R.; Fleissner,

    1993-01-01

    A new type of gamma-ray detector system for isotopic measurements has been developed. This new system, a ''Duo detector'' array, consists of two intrinsic germanium detectors, a planar followed by a coaxial mounted on the same axis within a single cryostat assembly. This configuration allows the isotopic analysis system to take advantage of spectral data results that are collected simultaneously from different gamma-ray energy regimes. Princeton Gamma Tech (PGT) produced several prototypes of this Duo detector array which were then tested by Rocky Flats personnel until the design was optimized. An application for this detector design is in automated, roboticized NDA systems such as those being developed at the Los Alamos TA-55 Plutonium Facility. The Duo detector design reduces the space necessary for the isotopic instrument by a factor of two (only one liquid nitrogen dewar is needed), and also reduces the complexity of the mechanical systems and controlling software. Data will be presented on measurements of nuclear material with a Duo detector for a wide variety of matrices. Results indicate that the maximum count rate can be increased up to 100,000 counts per second yet maintaining excellent resolution and energy rate product

  1. Robotic arm

    Science.gov (United States)

    Kwech, Horst

    1989-04-18

    A robotic arm positionable within a nuclear vessel by access through a small diameter opening and having a mounting tube supported within the vessel and mounting a plurality of arm sections for movement lengthwise of the mounting tube as well as for movement out of a window provided in the wall of the mounting tube. An end effector, such as a grinding head or welding element, at an operating end of the robotic arm, can be located and operated within the nuclear vessel through movement derived from six different axes of motion provided by mounting and drive connections between arm sections of the robotic arm. The movements are achieved by operation of remotely-controllable servo motors, all of which are mounted at a control end of the robotic arm to be outside the nuclear vessel.

  2. Overview of the aerial radiological measuring system (ARMS) program

    International Nuclear Information System (INIS)

    Deal, L.J.; Doyle, J.F. III.

    1975-01-01

    Since 1960 EG and G, Inc. has developed and maintained for the U.S. Energy Research and Development Administration (ERDA) a state-of-the-art radiation surveillance program called the Aerial Radiological Measuring System (ARMS). Radiological surveys covering more than 300,000 square miles have been performed. In addition to the radiation detector gear, the system includes an inertial navigation system, radar altimeter, meteorological probes, air sampler, air sample analyzer, multispectral cameras, aerial mapping camera and infrared scanner. The recently improved data acquisition system, REDAR, records all inputs digitally on magnetic tape and is readily mounted in fixed-wing aircraft or helicopters. The data analysis system, REDAC, is mounted in a mobile processing laboratory which accompanies the aircraft on surveys. Radiation isopleth maps, both for gross counts and selected isotopes, can be prepared in the field. Special computer software enables the ARMS to detect changes of less than 1.0 μR/hr in exposure rates between successive surveys of a given site

  3. Gas microstrip detectors on polymer, silicon and glass substrates

    International Nuclear Information System (INIS)

    Barasch, E.F.; Demroff, H.P.; Drew, M.M.; Elliott, T.S.; Gaedke, R.M.; Goss, L.T.; Kasprowicz, T.B.; Lee, B.; Mazumdar, T.K.; McIntyre, P.M.; Pang, Y.; Smith, D.D.; Trost, H.J.; Vanstraelen, G.; Wahl, J.

    1993-01-01

    We present results on the performance of Gas Microstrip Detectors on various substrates. These include a 300 μm anode-anode pitch pattern on Tempax borosilicate glass and ABS/copolyether, a 200 μm pattern on Upilex ''S'' polyimide, Texin 4215, Tedlar, ion-implanted Kapton, orientation-dependent etched flat-topped silicon (''knife-edge chamber''), and iron-vanadium glass, and a 100 μm pitch pattern on Upilex ''S'' and ion-implanted Kapton. (orig.)

  4. Robotic arm

    International Nuclear Information System (INIS)

    Kwech, H.

    1989-01-01

    A robotic arm positionable within a nuclear vessel by access through a small diameter opening and having a mounting tube supported within the vessel and mounting a plurality of arm sections for movement lengthwise of the mounting tube as well as for movement out of a window provided in the wall of the mounting tube is disclosed. An end effector, such as a grinding head or welding element, at an operating end of the robotic arm, can be located and operated within the nuclear vessel through movement derived from six different axes of motion provided by mounting and drive connections between arm sections of the robotic arm. The movements are achieved by operation of remotely-controllable servo motors, all of which are mounted at a control end of the robotic arm to be outside the nuclear vessel. 23 figs

  5. SU-E-I-53: Comparison of Kerma-Area-Product Between the Micro-Angiographic Fluoroscope (MAF) and a Flat Panel Detector (FPD) as Used in Neuro-Endovascular Procedures

    International Nuclear Information System (INIS)

    Vijayan, S; Rana, V; Nagesh, S Setlur; Xiong, Z; Rudin, S; Bednarek, D

    2015-01-01

    Purpose: To determine the reduction of integral dose to the patient when using the micro-angiographic fluoroscope (MAF) compared to when using the standard flat-panel detector (FPD) for the techniques used during neurointerventional procedures. Methods: The MAF is a small field-of-view, high resolution x-ray detector which captures 1024 x 1024 pixels with an effective pixel size of 35μm and is capable of real-time imaging up to 30 frames per second. The MAF was used in neuro-interventions during those parts of the procedure when high resolution was needed and the FPD was used otherwise. The technique parameters were recorded when each detector was used and the kerma-area-product (KAP) per image frame was determined. KAP values were calculated for seven neuro interventions using premeasured calibration files of output as a function of kVp and beam filtration and included the attenuation of the patient table for the frontal projections to be more representative of integral patient dose. The air kerma at the patient entrance was multiplied by the beam area at that point to obtain the KAP values. The ranges of KAP values per frame were determined for the range of technique parameters used during the clinical procedures. To appreciate the benefit of the higher MAF resolution in the region of interventional activity, DA technique parameters were generally used with the MAF. Results: The lowest and highest values of KAP per frame for the MAF in DA mode were 4 and 50 times lower, respectively, compared to those of the FPD in pulsed fluoroscopy mode. Conclusion: The MAF was used in those parts of the clinical procedures when high resolution and image quality was essential. The integral patient dose as represented by the KAP value was substantially lower when using the MAF than when using the FPD due to the much smaller volume of tissue irradiated. This research was supported in part by Toshiba Medical Systems Corporation and NIH Grant R01EB002873

  6. SU-E-I-53: Comparison of Kerma-Area-Product Between the Micro-Angiographic Fluoroscope (MAF) and a Flat Panel Detector (FPD) as Used in Neuro-Endovascular Procedures

    Energy Technology Data Exchange (ETDEWEB)

    Vijayan, S; Rana, V; Nagesh, S Setlur; Xiong, Z; Rudin, S; Bednarek, D [Toshiba Stroke and Vascular Research Center, University at Buffalo, Buffalo, NY (United States)

    2015-06-15

    Purpose: To determine the reduction of integral dose to the patient when using the micro-angiographic fluoroscope (MAF) compared to when using the standard flat-panel detector (FPD) for the techniques used during neurointerventional procedures. Methods: The MAF is a small field-of-view, high resolution x-ray detector which captures 1024 x 1024 pixels with an effective pixel size of 35μm and is capable of real-time imaging up to 30 frames per second. The MAF was used in neuro-interventions during those parts of the procedure when high resolution was needed and the FPD was used otherwise. The technique parameters were recorded when each detector was used and the kerma-area-product (KAP) per image frame was determined. KAP values were calculated for seven neuro interventions using premeasured calibration files of output as a function of kVp and beam filtration and included the attenuation of the patient table for the frontal projections to be more representative of integral patient dose. The air kerma at the patient entrance was multiplied by the beam area at that point to obtain the KAP values. The ranges of KAP values per frame were determined for the range of technique parameters used during the clinical procedures. To appreciate the benefit of the higher MAF resolution in the region of interventional activity, DA technique parameters were generally used with the MAF. Results: The lowest and highest values of KAP per frame for the MAF in DA mode were 4 and 50 times lower, respectively, compared to those of the FPD in pulsed fluoroscopy mode. Conclusion: The MAF was used in those parts of the clinical procedures when high resolution and image quality was essential. The integral patient dose as represented by the KAP value was substantially lower when using the MAF than when using the FPD due to the much smaller volume of tissue irradiated. This research was supported in part by Toshiba Medical Systems Corporation and NIH Grant R01EB002873.

  7. [Measuring the blood pressure in both arms is of little use; longitudinal study into blood pressure differences between both arms and its reproducibility in patients with diabetes mellitus type 2].

    Science.gov (United States)

    Kleefstra, N; Houweling, S T; Meyboom-de Jong, B; Bilo, H J G

    2007-07-07

    To determine the prevalence of inter-arm blood pressure differences > 10 mmHg in patients with diabetes mellitus type 2 (DM2) and to determine whether these differences are consistent over time. Descriptive. In an evaluation study of 169 DM2 patients from 5 general practices in 2003 and 2004, different methods of oscillatory measurement were used to investigate inter-arm blood pressure differences > 10 mmHg systolic or diastolic. These methods were: one measurement in each arm non-simultaneously (method A), one measurement simultaneously (B) and the mean of two simultaneous measurements (C). With method A an inter-arm blood pressure difference was found in 33% of patients. This percentage diminished to 9 with method C. In 44% (n = 7) of the patients in whom method C detected a relevant blood pressure difference, this difference was not found with method A. In 79% of patients the inter-arm blood pressure difference was not reproduced after one year. In daily practice, one non-simultaneous blood pressure measurement in each arm (method A) was of little value for identification of patients with inter-arm blood pressure differences. The reproducibility was poor one year later. Bilateral blood pressure measurement is therefore of little value.

  8. Near-field flat focusing mirrors

    Science.gov (United States)

    Cheng, Yu-Chieh; Staliunas, Kestutis

    2018-03-01

    This article reviews recent progress towards the design of near-field flat focusing mirrors, focusing/imaging light patterns in reflection. An important feature of such flat focusing mirrors is their transverse invariance, as they do not possess any optical axis. We start with a review of the physical background to the different focusing mechanisms of near- and far-field focusing. These near-field focusing devices like flat lenses and the reviewed near-field focusing mirrors can implement planar focusing devices without any optical axis. In contrast, various types of far-field planar focusing devices, such as high-contrast gratings and metasurfaces, unavoidably break the transverse invariance due to their radially symmetrical structures. The particular realizations of near-field flat focusing mirrors including Bragg-like dielectric mirrors and dielectric subwavelength gratings are the main subjects of the review. The first flat focusing mirror was demonstrated with a chirped mirror and was shown to manage an angular dispersion for beam focusing, similar to the management of chromatic dispersion for pulse compression. Furthermore, the reviewed optimized chirped mirror demonstrated a long near-field focal length, hardly achieved by a flat lens or a planar hyperlens. Two more different configurations of dielectric subwavelength gratings that focus a light beam at normal or oblique incidence are also reviewed. We also summarize and compare focusing performance, limitations, and future perspectives between the reviewed flat focusing mirrors and other planar focusing devices including a flat lens with a negative-index material, a planar hyperlens, a high-contrast grating, and a metasurface.

  9. Position Control Method For Pick And Place Robot Arm For Object Sorting System

    Directory of Open Access Journals (Sweden)

    Khin Moe Myint

    2015-08-01

    Full Text Available The more increase the number of industries in developing countries the more require labourers or workers in that. To reduce the cost of labour force and to increase the manufacturing capacity of industries the advanced robot arms are more needed. The aim of this journal is to eliminate the manual control for object sorting system.Robot arm design in this research uses two joints three links and servo motors to drive. Microcontroller is used to generate required PWM signal for servo motors. In this research the position control of robot arm was designed by using kinematic control methods. There are two types of kinematic control methods which are forward and reverse kinematic methods. In forward kinematic method the input parameters are the joint angles and link length of robot arm and then the output is the position at XYZ coordinate of tool or gripper. In inverse kinematic the input parameters are position at XYZ coordinate of gripper and the link length of robot arm and then the output parameters are the joint angles. So kinematic methods can explain the analytical description of the geometry motion of the manipulator with reference to a robot coordinate system fixed to a frame without consideration of the forces or the moments causing the movements. For sorting system Metal detector is used to detect the metal or non-metal. This position control of pick and place robot arm is fully tested and the result is obtained more precisely.

  10. SUPERCOLLIDER: A GEM of a detector

    International Nuclear Information System (INIS)

    Anon.

    1993-01-01

    Now being prepared as a major experimental facility for the 87- kilometre Superconducting Supercollider (SSC) being built in Ellis County, Texas, is the GEM detector project. GEM thus becomes the companion to the Solenoidal Detector Collaboration (SDC), the first major SSC detector to emerge (March 1992, page 13). This is in keeping with the SSC Laboratory's aim of two major detectors with overlapping and complementary strengths. GEM is designed to observe all SSC signatures, with emphasis on precise measurement of electrons, photons and muons. Hence the name GEM - ''Gammas, Electrons and Muons.'' Design goals are clean signatures for leptons, jets, and missing transverse energy, maximum sensitivity to narrow resonances, and low backgrounds. Also important is maintaining significant capability at high luminosity (10 34 cm -2 s -1 ). GEM has some distinctive features. A key concept is the exterior magnet, surrounding all detector elements. Inside the magnet are a muon tracking system, a precision calorimeter, and a compact central tracker. This allows the muon momentum to be measured the air of the radiation shielded area outside the thick calorimeter, giving both high precision and robustness at high luminosity. A large magnet gives a large lever arm (at least 4 m) for precise muon momentum measurement. Placing the magnet outside also minimizes the material between tracker and calorimeters, so that the calorimeters are limited only by their inherent resolutions

  11. Performance evaluation of a direct-conversion flat-panel detector system in imaging and quality assurance for a high-dose-rate 192Ir source

    Science.gov (United States)

    Miyahara, Yoshinori; Hara, Yuki; Nakashima, Hiroto; Nishimura, Tomonori; Itakura, Kanae; Inomata, Taisuke; Kitagaki, Hajime

    2018-03-01

    In high-dose-rate (HDR) brachytherapy, a direct-conversion flat-panel detector (d-FPD) clearly depicts a 192Ir source without image halation, even under the emission of high-energy gamma rays. However, it was unknown why iridium is visible when using a d-FPD. The purpose of this study was to clarify the reasons for visibility of the source core based on physical imaging characteristics, including the modulation transfer functions (MTF), noise power spectral (NPS), contrast transfer functions, and linearity of d-FPD to high-energy gamma rays. The acquired data included: x-rays, [X]; gamma rays, [γ] dual rays (X  +  γ), [D], and subtracted data for depicting the source ([D]  -  [γ]). In the quality assurance (QA) test for the positional accuracy of a source core, the coordinates of each dwelling point were compared between the planned and actual source core positions using a CT/MR-compatible ovoid applicator and a Fletcher-Williamson applicator. The profile curves of [X] and ([D]  -  [γ]) matched well on MTF and NPS. The contrast resolutions of [D] and [X] were equivalent. A strongly positive linear correlation was found between the output data of [γ] and source strength (r 2  >  0.99). With regard to the accuracy of the source core position, the largest coordinate difference (3D distance) was noted at the maximum curvature of the CT/MR-compatible ovoid and Fletcher-Williamson applicators, showing 1.74  ±  0.02 mm and 1.01  ±  0.01 mm, respectively. A d-FPD system provides high-quality images of a source, even when high-energy gamma rays are emitted to the detector, and positional accuracy tests with clinical applicators are useful in identifying source positions (source movements) within the applicator for QA.

  12. Blowup for flat slow manifolds

    DEFF Research Database (Denmark)

    Kristiansen, Kristian Uldall

    2017-01-01

    In this paper, we present a way of extending the blowup method, in the formulation of Krupa and Szmolyan, to flat slow manifolds that lose hyperbolicity beyond any algebraic order. Although these manifolds have infinite co-dimensions, they do appear naturally in certain settings; for example, in (a......) the regularization of piecewise smooth systems by tanh, (b) a particular aircraft landing dynamics model, and finally (c) in a model of earthquake faulting. We demonstrate the approach using a simple model system and the examples (a) and (b)....

  13. Blowup for flat slow manifolds

    Science.gov (United States)

    Kristiansen, K. U.

    2017-05-01

    In this paper, we present a way of extending the blowup method, in the formulation of Krupa and Szmolyan, to flat slow manifolds that lose hyperbolicity beyond any algebraic order. Although these manifolds have infinite co-dimensions, they do appear naturally in certain settings; for example, in (a) the regularization of piecewise smooth systems by \\tanh , (b) a particular aircraft landing dynamics model, and finally (c) in a model of earthquake faulting. We demonstrate the approach using a simple model system and the examples (a) and (b).

  14. The development of a general purpose ARM-based processing unit for the ATLAS TileCal sROD

    Science.gov (United States)

    Cox, M. A.; Reed, R.; Mellado, B.

    2015-01-01

    After Phase-II upgrades in 2022, the data output from the LHC ATLAS Tile Calorimeter will increase significantly. ARM processors are common in mobile devices due to their low cost, low energy consumption and high performance. It is proposed that a cost-effective, high data throughput Processing Unit (PU) can be developed by using several consumer ARM processors in a cluster configuration to allow aggregated processing performance and data throughput while maintaining minimal software design difficulty for the end-user. This PU could be used for a variety of high-level functions on the high-throughput raw data such as spectral analysis and histograms to detect possible issues in the detector at a low level. High-throughput I/O interfaces are not typical in consumer ARM System on Chips but high data throughput capabilities are feasible via the novel use of PCI-Express as the I/O interface to the ARM processors. An overview of the PU is given and the results for performance and throughput testing of four different ARM Cortex System on Chips are presented.

  15. The development of a general purpose ARM-based processing unit for the ATLAS TileCal sROD

    International Nuclear Information System (INIS)

    Cox, M A; Reed, R; Mellado, B

    2015-01-01

    After Phase-II upgrades in 2022, the data output from the LHC ATLAS Tile Calorimeter will increase significantly. ARM processors are common in mobile devices due to their low cost, low energy consumption and high performance. It is proposed that a cost-effective, high data throughput Processing Unit (PU) can be developed by using several consumer ARM processors in a cluster configuration to allow aggregated processing performance and data throughput while maintaining minimal software design difficulty for the end-user. This PU could be used for a variety of high-level functions on the high-throughput raw data such as spectral analysis and histograms to detect possible issues in the detector at a low level. High-throughput I/O interfaces are not typical in consumer ARM System on Chips but high data throughput capabilities are feasible via the novel use of PCI-Express as the I/O interface to the ARM processors. An overview of the PU is given and the results for performance and throughput testing of four different ARM Cortex System on Chips are presented

  16. Burnout detector design for heat transfer experiments

    International Nuclear Information System (INIS)

    Dias, H.F.

    1992-01-01

    This paper describes the design of an burnout detector for heat transfer experiments, applied during tests for optimization of fuel elements for PWR reactors. The burnout detector avoids the fuel rods destruction during the experiments at the Centro de Desenvolvimento da Tecnologia Nuclear. The detector evaluates the temperature changes over the fuel rods in the temperature changes over the fuel rods in the area where the burnout phenomenon could be anticipated. As soon as the phenomenon appears, the system power supply is turned off. The thermal Circuit No. 1, during the experiments, had been composed by nine fuel rods feed parallelly by the same power supply. Fine copper wires had been attached at the centre and at the ends of the fuel rod to take two Wheat stone bridge arms. The detector had been applied across the bridge diagonals, which must be balanced the burnout excursion can be detected as a small but fast increase of the signal over the detector. Large scale experiments had been carried out to compare the resistance bridge performance against a thermocouple attached through the fuel rod wall. These experiments had been showed us the advantages of the first method over the last, because the bridge evaluates the whole fuel rod, while the thermocouple evaluates only the area where it had been attached. (author)

  17. The spiral arms of the Milky Way: The relative location of each different arm tracer within a typical spiral arm width

    Energy Technology Data Exchange (ETDEWEB)

    Vallée, Jacques P., E-mail: jacques.vallee@nrc-cnrc.gc.ca [National Research Council Canada, National Science Infrastructure portfolio, Herzberg Astronomy and Astrophysics, 5071 West Saanich Road, Victoria, B.C., V9E 2E7 (Canada)

    2014-07-01

    From the Sun's location in the Galactic disk, different arm tracers (CO, H I, hot dust, etc.) have been employed to locate a tangent to each spiral arm. Using all various and different observed spiral arm tracers (as published elsewhere), we embark on a new goal, namely the statistical analysis of these published data (data mining) to statistically compute the mean location of each spiral arm tracer. We show for a typical arm cross-cut, a separation of 400 pc between the mid-arm and the dust lane (at the inner edge of the arm, toward the Galactic center). Are some arms major and others minor? Separating arms into two sets, as suggested by some, we find the same arm widths between the two sets. Our interpretation is that we live in a multiple (four-arm) spiral (logarithmic) pattern (around a pitch angle of 12°) for the stars and gas in the Milky Way, with a sizable interarm separation (around 3 kpc) at the Sun's location and the same arm width for each arm (near 400 pc from mid-arm to dust lane).

  18. The spiral arms of the Milky Way: The relative location of each different arm tracer within a typical spiral arm width

    International Nuclear Information System (INIS)

    Vallée, Jacques P.

    2014-01-01

    From the Sun's location in the Galactic disk, different arm tracers (CO, H I, hot dust, etc.) have been employed to locate a tangent to each spiral arm. Using all various and different observed spiral arm tracers (as published elsewhere), we embark on a new goal, namely the statistical analysis of these published data (data mining) to statistically compute the mean location of each spiral arm tracer. We show for a typical arm cross-cut, a separation of 400 pc between the mid-arm and the dust lane (at the inner edge of the arm, toward the Galactic center). Are some arms major and others minor? Separating arms into two sets, as suggested by some, we find the same arm widths between the two sets. Our interpretation is that we live in a multiple (four-arm) spiral (logarithmic) pattern (around a pitch angle of 12°) for the stars and gas in the Milky Way, with a sizable interarm separation (around 3 kpc) at the Sun's location and the same arm width for each arm (near 400 pc from mid-arm to dust lane).

  19. Two-arm semiconductor spectrometer for charged particles for the investigation of absorption by nuclei of stopped negative pions

    International Nuclear Information System (INIS)

    Gornov, M.G.; Gurov, Yu.B.; Lukin, A.S.; Morokhov, P.V.; Pechkurov, V.A.; Pichugin, A.P.; Saveliev, V.I.; Shafigullin, R.R.; Sergeev, F.M.; Khomutov, A.A.; Oganesian, K.O.; Osipenko, B.P.; Sandukovsky, V.G.

    1984-01-01

    A two-arm semiconductor spectrometer for the detection of secondary charged particles, such as p, d, t, 3 He, 4 He and of their correlations in the process of stopped pion absorption by nuclei is described. The spectrometer consists of two telescopes of Si-detectors with a sensitive surface of 800 mm 2 , two semiconductor detectors as monitors and the 'live' target, a silicon surface barrier detector. The number of semiconductor detectors is 19. A technique for pion stop selection in thin targets is described. The problem of particle identification and of measurements of their energy with the help of a multicrystal semiconductor telescope is discussed. The technique provides an absolute normalization of spectra. The 'alive' target helps to obtain more information on the process of pion absorption by Si-nuclei. (orig.)

  20. Low-bias flat band-stop filter based on velocity modulated gaussian graphene superlattice

    Science.gov (United States)

    Sattari-Esfahlan, S. M.; Shojaei, S.

    2018-05-01

    Transport properties of biased planar Gaussian graphene superlattice (PGGSL) with Fermi velocity barrier is investigated by transfer matrix method (TMM). It is observed that enlargement of bias voltage over miniband width breaks the miniband to WSLs leads to suppressing resonant tunneling. Transmission spectrum shows flat wide stop-band property controllable by external bias voltage with stop-band width of near 200 meV. The simulations demonstrate that strong velocity barriers prevent tunneling of Dirac electrons leading to controllable enhancement of stop-band width. By increasing ratio of Fermi velocity in barriers to wells υc stop-band width increase. As wide transmission stop-band width (BWT) of filter is tunable from 40 meV to 340 meV is obtained by enhancing ratio of υc from 0.2 to 1.5, respectively. Proposed structure suggests easy tunable wide band-stop electronic filter with a modulated flat stop-band characteristic by height of electrostatic barrier and structural parameters. Robust sensitivity of band width to velocity barrier intensity in certain bias voltages and flat band feature of proposed filter may be opens novel venue in GSL based flat band low noise filters and velocity modulation devices.

  1. Compensating for Channel Fading in DS-CDMA Communication Systems Employing ICA Neural Network Detectors

    Directory of Open Access Journals (Sweden)

    David Overbye

    2005-06-01

    Full Text Available In this paper we examine the impact of channel fading on the bit error rate of a DS-CDMA communication system. The system employs detectors that incorporate neural networks effecting methods of independent component analysis (ICA, subspace estimation of channel noise, and Hopfield type neural networks. The Rayleigh fading channel model is used. When employed in a Rayleigh fading environment, the ICA neural network detectors that give superior performance in a flat fading channel did not retain this superior performance. We then present a new method of compensating for channel fading based on the incorporation of priors in the ICA neural network learning algorithms. When the ICA neural network detectors were compensated using the incorporation of priors, they give significantly better performance than the traditional detectors and the uncompensated ICA detectors. Keywords: CDMA, Multi-user Detection, Rayleigh Fading, Multipath Detection, Independent Component Analysis, Prior Probability Hebbian Learning, Natural Gradient

  2. Highly sensitive x-ray detectors in the low-energy range on n-type 4H-SiC epitaxial layers

    Energy Technology Data Exchange (ETDEWEB)

    Mandal, Krishna C.; Muzykov, Peter G. [Department of Electrical Engineering, University of South Carolina, Columbia, South Carolina 29208 (United States); Russell Terry, J. [Space Science and Applications Group (ISR-1), Intelligence and Space Research Division, Los Alamos National Laboratory, Los Alamos, New Mexico 87545 (United States)

    2012-07-30

    Schottky diodes on n-type 4H-SiC epitaxial layers have been fabricated for low-energy x-ray detection. The detectors were highly sensitive to soft x-rays and showed improved response compared to the commercial SiC UV photodiodes. Current-voltage characteristics at 475 K showed low leakage current revealing the possibility of high temperature operation. The high quality of the epi-layer was confirmed by x-ray diffraction and chemical etching. Thermally stimulated current measurements performed at 94-550 K revealed low density of deep levels which may cause charge trapping. No charge trapping on detectors' responsivity in the low x-ray energy was found.

  3. Physical characteristics of GE Senographe Essential and DS digital mammography detectors

    International Nuclear Information System (INIS)

    Ghetti, Caterina; Borrini, Adriano; Ortenzia, Ornella; Rossi, Raffaella; Ordonez, Pedro L.

    2008-01-01

    The purpose of this study was to investigate physical characteristics of two full field digital mammography (FFDM) systems (GE Senographe Essential and DS). Both are indirect conversion (x ray to light) a-Si flat panels coupled with a CsI(Tl) scintillator. The examined systems have the same pixel size (100 μm) but a different field of view: a conventional size 23x19.2 cm 2 and a large field 24x30.7 cm 2 , specifically designed to image large breasts. In the GE Senographe Essential model relevant improvements in flat panel design were implemented and new deposition tools for metal, a-Si, and CsI(Tl) were introduced by GE. These changes in detector design are expected to be beneficial for advanced applications such as breast tomosynthesis. The presampling modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) were measured for a wide range of exposure (25-240 μGy) with a RQA-M2 technique (28 kVp with a Mo/Mo target/filter combination and 2 mm of additional aluminum filtration). At 1, 2, and at 4 lp/mm MTF is equal to 0.9, 0.76, and 0.46 for the conventional field detector and to 0.85, 0.59, and 0.24 for the large field detector. The latter detector exhibits an improved NNPS due to a lower electronic noise and a better DQE that reaches 60%. In addition a contrast-detail analysis was performed with CDMAM 3.4 phantom and CDCOM software: GE Senographe DS showed statistically significant poorer detection ability in comparison with the GE Senographe Essential. These results could have been expected, at least qualitatively, considering the relative DQE of the two systems

  4. Low Energy X-Ray and γ-Ray Detectors Fabricated on n-Type 4H-SiC Epitaxial Layer

    Science.gov (United States)

    Mandal, Krishna C.; Muzykov, Peter G.; Chaudhuri, Sandeep K.; Terry, J. Russell

    2013-08-01

    Schottky barrier diode (SBD) radiation detectors have been fabricated on n-type 4H-SiC epitaxial layers and evaluated for low energy x- and γ-rays detection. The detectors were found to be highly sensitive to soft x-rays in the 50 eV to few keV range and showed 2.1 % energy resolution for 59.6 keV gamma rays. The response to soft x-rays for these detectors was significantly higher than that of commercial off-the-shelf (COTS) SiC UV photodiodes. The devices have been characterized by current-voltage (I-V) measurements in the 94-700 K range, thermally stimulated current (TSC) spectroscopy, x-ray diffraction (XRD) rocking curve measurements, and defect delineating chemical etching. I-V characteristics of the detectors at 500 K showed low leakage current ( nA at 200 V) revealing a possibility of high temperature operation. The XRD rocking curve measurements revealed high quality of the epitaxial layer exhibiting a full width at half maximum (FWHM) of the rocking curve 3.6 arc sec. TSC studies in a wide range of temperature (94-550 K) revealed presence of relatively shallow levels ( 0.25 eV) in the epi bulk with a density 7×1013 cm-3 related to Al and B impurities and deeper levels located near the metal-semiconductor interface.

  5. Assembly and evaluation of a pyroelectric detector bonded to vertically aligned multiwalled carbon nanotubes over thin silicon.

    Science.gov (United States)

    Theocharous, E; Theocharous, S P; Lehman, J H

    2013-11-20

    A novel pyroelectric detector consisting of a vertically aligned nanotube array on thin silicon (VANTA/Si) bonded to a 60 μm thick crystal of LiTaO₃ has been fabricated. The performance of the VANTA/Si-coated pyroelectric detector was evaluated using National Physical Laboratory's (NPL's) detector-characterization facilities. The relative spectral responsivity of the detector was found to be spectrally flat in the 0.8-24 μm wavelength range, in agreement with directional-hemispherical reflectance measurements of witness samples of the VANTA. The spatial uniformity of response of the test detector exhibited good uniformity, although the nonuniformity increased with increasing modulation frequency. The nonuniformity may be assigned either to the dimensions of the VANTA or the continuity of the bond between the VANTA/Si coating and the pyroelectric crystal substrate. The test detector exhibited a small superlinear response, which is similar to that of pyroelectric detectors coated with good quality gold-black coatings.

  6. 75 FR 55745 - Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Preliminary Results...

    Science.gov (United States)

    2010-09-14

    ... Products covered by this order are certain corrosion-resistant carbon steel flat products from Korea. These... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon... review of the countervailing duty (CVD) order on corrosion-resistant carbon steel flat products (CORE...

  7. Angular-momentum-dominated electron beams and flat-beam generation

    International Nuclear Information System (INIS)

    Sun, Yin-e

    2005-01-01

    In the absence of external forces, if the dynamics within an electron beam is dominated by its angular momentum rather than other effects such as random thermal motion or self Coulomb-repulsive force (i.e., space-charge force), the beam is said to be angular-momentum-dominated. Such a beam can be directly applied to the field of electron-cooling of heavy ions; or it can be manipulated into an electron beam with large transverse emittance ratio, i.e., a flat beam. A flat beam is of interest for high-energy electron-positron colliders or accelerator-based light sources. An angular-momentum-dominated beam is generated at the Fermilab/NICADD photoinjector Laboratory (FNPL) and is accelerated to an energy of 16 MeV. The properties of such a beam is investigated systematically in experiment. The experimental results are in very good agreement with analytical expectations and simulation results. This lays a good foundation for the transformation of an angular-momentum-dominated beam into a flat beam. The round-to-flat beam transformer is composed of three skew quadrupoles. Based on a good knowledge of the angular-momentum-dominated beam, the quadrupoles are set to the proper strengths in order to apply a total torque which removes the angular momentum, resulting in a flat beam. For bunch charge around 0.5 nC, an emittance ratio of 100 ± 5 was measured, with the smaller normalized root-mean-square emittance around 0.4 mm-mrad. Effects limiting the flat-beam emittance ratio are investigated, such as the chromatic effects in the round-to-flat beam transformer, asymmetry in the initial angular-momentum-dominated beam, and space-charge effects. The most important limiting factor turns out to be the uncorrelated emittance growth caused by space charge when the beam energy is low, for example, in the rf gun area. As a result of such emittance growth prior to the round-to-flat beam transformer, the emittance ratio achievable in simulation decreases from orders of thousands to

  8. Angular-momentum-dominated electron beams and flat-beam generation

    Energy Technology Data Exchange (ETDEWEB)

    Sun, Yin-e [Univ. of Chicago, IL (United States)

    2005-06-01

    In the absence of external forces, if the dynamics within an electron beam is dominated by its angular momentum rather than other effects such as random thermal motion or self Coulomb-repulsive force (i.e., space-charge force), the beam is said to be angular-momentum-dominated. Such a beam can be directly applied to the field of electron-cooling of heavy ions; or it can be manipulated into an electron beam with large transverse emittance ratio, i.e., a flat beam. A flat beam is of interest for high-energy electron-positron colliders or accelerator-based light sources. An angular-momentum-dominated beam is generated at the Fermilab/NICADD photoinjector Laboratory (FNPL) and is accelerated to an energy of 16 MeV. The properties of such a beam is investigated systematically in experiment. The experimental results are in very good agreement with analytical expectations and simulation results. This lays a good foundation for the transformation of an angular-momentum-dominated beam into a flat beam. The round-to-flat beam transformer is composed of three skew quadrupoles. Based on a good knowledge of the angular-momentum-dominated beam, the quadrupoles are set to the proper strengths in order to apply a total torque which removes the angular momentum, resulting in a flat beam. For bunch charge around 0.5 nC, an emittance ratio of 100 ± 5 was measured, with the smaller normalized root-mean-square emittance around 0.4 mm-mrad. Effects limiting the flat-beam emittance ratio are investigated, such as the chromatic effects in the round-to-flat beam transformer, asymmetry in the initial angular-momentum-dominated beam, and space-charge effects. The most important limiting factor turns out to be the uncorrelated emittance growth caused by space charge when the beam energy is low, for example, in the rf gun area. As a result of such emittance growth prior to the round-to-flat beam transformer, the emittance ratio achievable in simulation decreases from orders of thousands to

  9. OP09O-OP404-9 Wide Field Camera 3 CCD Quantum Efficiency Hysteresis

    Science.gov (United States)

    Collins, Nick

    2009-01-01

    The HST/Wide Field Camera (WFC) 3 UV/visible channel CCD detectors have exhibited an unanticipated quantum efficiency hysteresis (QEH) behavior. At the nominal operating temperature of -83C, the QEH feature contrast was typically 0.1-0.2% or less. The behavior was replicated using flight spare detectors. A visible light flat-field (540nm) with a several times full-well signal level can pin the detectors at both optical (600nm) and near-UV (230nm) wavelengths, suppressing the QEH behavior. We are characterizing the timescale for the detectors to become unpinned and developing a protocol for flashing the WFC3 CCDs with the instrument's internal calibration system in flight. The HST/Wide Field Camera 3 UV/visible channel CCD detectors have exhibited an unanticipated quantum efficiency hysteresis (QEH) behavior. The first observed manifestation of QEH was the presence in a small percentage of flat-field images of a bowtie-shaped contrast that spanned the width of each chip. At the nominal operating temperature of -83C, the contrast observed for this feature was typically 0.1-0.2% or less, though at warmer temperatures contrasts up to 5% (at -50C) have been observed. The bowtie morphology was replicated using flight spare detectors in tests at the GSFC Detector Characterization Laboratory by power cycling the detector while cold. Continued investigation revealed that a clearly-related global QE suppression at the approximately 5% level can be produced by cooling the detector in the dark; subsequent flat-field exposures at a constant illumination show asymptotically increasing response. This QE "pinning" can be achieved with a single high signal flat-field or a series of lower signal flats; a visible light (500-580nm) flat-field with a signal level of several hundred thousand electrons per pixel is sufficient for QE pinning at both optical (600nm) and near-UV (230nm) wavelengths. We are characterizing the timescale for the detectors to become unpinned and developing a

  10. Energetic ion loss detector on the Alcator C-Mod tokamak.

    Science.gov (United States)

    Pace, D C; Granetz, R S; Vieira, R; Bader, A; Bosco, J; Darrow, D S; Fiore, C; Irby, J; Parker, R R; Parkin, W; Reinke, M L; Terry, J L; Wolfe, S M; Wukitch, S J; Zweben, S J

    2012-07-01

    A scintillator-based energetic ion loss detector has been successfully commissioned on the Alcator C-Mod tokamak. This probe is located just below the outer midplane, where it captures ions of energies up to 2 MeV resulting from ion cyclotron resonance heating. After passing through a collimating aperture, ions impact different regions of the scintillator according to their gyroradius (energy) and pitch angle. The probe geometry and installation location are determined based on modeling of expected lost ions. The resulting probe is compact and resembles a standard plasma facing tile. Four separate fiber optic cables view different regions of the scintillator to provide phase space resolution. Evolving loss levels are measured during ion cyclotron resonance heating, including variation dependent upon individual antennae.

  11. One-pot synthesis of star-shaped macromolecules containing polyglycidol and poly(ethylene oxide) arms.

    Science.gov (United States)

    Lapienis, Grzegorz; Penczek, Stanislaw

    2005-01-01

    Synthesis of fully hydrophilic star-shaped macromolecules with different kinds of arms (A(x)B(y)C(z)) based on polyglycidol (PGL, A(x)) and poly(ethylene oxide) (PEO, C(z)) arms and diepoxy compounds (diglycidyl ethers of ethylene glycol (DGEG) or neopentyl glycol (DGNG) in the core, B(y)) forming the core is described. Precursors of arms were prepared by polymerization of glycidol with protected -OH groups. The first-generation stars were formed in the series of consecutive-parallel reactions of arms A(x) with diepoxy compounds (B). These first-generation stars (A(x)B(y)), having approximately O-, Mt+ groups on the cores, were used as multianionic initiators for the second generation of arms (C(z)) built by polymerization of ethylene oxide. The products with M(n) up to 10(5) and having up to approximately 40 arms were obtained. The number of arms (f) was determined by direct measurements of M(n) of the first-generation stars (M(n) of arms A(x) is known), compared with f calculated from the branching index g, determined from R(g) measured with size-exclusion chromatography (SEC) triple detection with TriSEC software. The progress of the star formation was monitored by 1H NMR and SEC. These novel water-soluble stars, having a large number of hydroxyl groups, both at the ends of PEO arms as well as within the PGL arms, can be functionalized and further used for attaching compounds of interest. This approach opens, therefore, a new way of "multiPEGylation".

  12. Space-based detectors

    DEFF Research Database (Denmark)

    Sesana, A.; Weber, W. J.; Killow, C. J.

    2014-01-01

    ) is planned for 2015. This mission and its payload “LISA Technology Package” will demonstrate key technologies for LISA. In this context, reference masses in free fall for LISA, and gravitational physics in general, was described by William Weber, laser interferometry at the pico-metre level and the optical...... of the LISA technology that are not going to be demonstrated by LPF, but under intensive development at the moment, were presented by Oliver Jennrich and Oliver Gerberding. Looking into the future, Japan is studying the design of a mid-frequency detector called DECIGO, which was discussed by Tomotada Akutsu...... as technology demonstrator. This will be the first inter-spacecraft laser interferometer and has many aspects in common with the LISA long arm, as discussed by Andrew Sutton....

  13. Novel fabrication techniques for low-mass composite structures in silicon particle detectors

    Energy Technology Data Exchange (ETDEWEB)

    Hartman, Neal, E-mail: neal.hartman@cern.ch; Silber, Joseph; Anderssen, Eric; Garcia-Sciveres, Maurice; Gilchriese, Murdock; Johnson, Thomas; Cepeda, Mario

    2013-12-21

    The structural design of silicon-based particle detectors is governed by competing demands of reducing mass while maximizing stability and accuracy. These demands can only be met by fiber reinforced composite laminates (CFRP). As detecting sensors and electronics become lower mass, the motivation to reduce structure as a proportion of overall mass pushes modern detector structures to the lower limits of composite ply thickness, while demanding maximum stiffness. However, classical approaches to composite laminate design require symmetric laminates and flat structures, in order to minimize warping during fabrication. This constraint of symmetry in laminate design, and a “flat plate” approach to fabrication, results in more massive structures. This study presents an approach to fabricating stable and accurate, geometrically complex composite structures by bonding warped, asymmetric, but ultra-thin component laminates together in an accurate tool, achieving final overall precision normally associated with planar structures. This technique has been used to fabricate a prototype “I-beam” that supports two layers of detecting elements, while being up to 20 times stiffer and up to 30% lower mass than comparable, independent planar structures (typically known as “staves”)

  14. Development and application of W/Cu flat-type plasma facing components at ASIPP

    Science.gov (United States)

    Li, Q.; Zhao, S. X.; Sun, Z. X.; Xu, Y.; Li, B.; Wei, R.; Wang, W. J.; Qin, S. G.; Shi, Y. L.; Xie, C. Y.; Wang, J. C.; Wang, X. L.; Missirlian, M.; Guilhem, D.; Liu, G. H.; Yang, Z. S.; Luo, G.-N.

    2017-12-01

    W/Cu flat-type plasma facing components (PFCs) were widely used in divertor of fusion device because of its advantages, such as low cost, light in weight and good machinability. However, it is very difficult to manufacture them due to the large mismatch between the thermo-mechanical properties of W and Cu. Institute of Plasma Physics, Chinese Academy of Sciences (ASIPP) has successfully developed W/Cu flat-type PFCs for EAST W/Cu divertor project by hot isostatic pressing (HIP) technology. This paper presents the development and application of W/Cu flat-type PFCs at ASIPP. The optimized manufacturing process is to cast pure copper onto the rear side of W tiles at temperature of 1200 °C firstly, and then to HIP the W/Cu tiles onto CuCrZr heat sink at temperature of 600 °C, pressure of 150 MPa and duration of 3 h. W/Cu flat-type testing mock-up for EAST survived 1000 cycles at heat load of 5 MW m-2 in high heat flux tests. And then ASIPP prepared two mock-ups for CEA’s tungsten environment in steady-state tokamak (WEST) project. One mock-up withstood successfully 302 cycles of 20 MW m-2, which are far beyond the design requirement. Since 2014, W/Cu flat-type PFCs were wildly used in EAST upper divertor as baffle and dome components which showed excellent performance in 2015 and 2016 campaigns. Given the success in EAST upper divertor, W/Cu flat-type concept is as well applied in the design of actively cooled Langmuir probes which will be mounted onto EAST divertor targets soon.

  15. Development and application of W/Cu flat-type plasma facing components at ASIPP

    International Nuclear Information System (INIS)

    Li, Q; Sun, Z X; Xu, Y; Li, B; Wei, R; Wang, W J; Xie, C Y; Wang, J C; Wang, X L; Yang, Z S; Luo, G-N; Zhao, S X; Qin, S G; Shi, Y L; Liu, G H; Missirlian, M; Guilhem, D

    2017-01-01

    W/Cu flat-type plasma facing components (PFCs) were widely used in divertor of fusion device because of its advantages, such as low cost, light in weight and good machinability. However, it is very difficult to manufacture them due to the large mismatch between the thermo-mechanical properties of W and Cu. Institute of Plasma Physics, Chinese Academy of Sciences (ASIPP) has successfully developed W/Cu flat-type PFCs for EAST W/Cu divertor project by hot isostatic pressing (HIP) technology. This paper presents the development and application of W/Cu flat-type PFCs at ASIPP. The optimized manufacturing process is to cast pure copper onto the rear side of W tiles at temperature of 1200 °C firstly, and then to HIP the W/Cu tiles onto CuCrZr heat sink at temperature of 600 °C, pressure of 150 MPa and duration of 3 h. W/Cu flat-type testing mock-up for EAST survived 1000 cycles at heat load of 5 MW m −2 in high heat flux tests. And then ASIPP prepared two mock-ups for CEA’s tungsten environment in steady-state tokamak (WEST) project. One mock-up withstood successfully 302 cycles of 20 MW m −2 , which are far beyond the design requirement. Since 2014, W/Cu flat-type PFCs were wildly used in EAST upper divertor as baffle and dome components which showed excellent performance in 2015 and 2016 campaigns. Given the success in EAST upper divertor, W/Cu flat-type concept is as well applied in the design of actively cooled Langmuir probes which will be mounted onto EAST divertor targets soon. (paper)

  16. A novel high resolution, high sensitivity SPECT detector for molecular imaging of cardiovascular diseases

    Science.gov (United States)

    Cusanno, F.; Argentieri, A.; Baiocchi, M.; Colilli, S.; Cisbani, E.; De Vincentis, G.; Fratoni, R.; Garibaldi, F.; Giuliani, F.; Gricia, M.; Lucentini, M.; Magliozzi, M. L.; Majewski, S.; Marano, G.; Musico, P.; Musumeci, M.; Santavenere, F.; Torrioli, S.; Tsui, B. M. W.; Vitelli, L.; Wang, Y.

    2010-05-01

    Cardiovascular diseases are the most common cause of death in western countries. Understanding the rupture of vulnerable atherosclerotic plaques and monitoring the effect of innovative therapies of heart failure is of fundamental importance. A flexible, high resolution, high sensitivity detector system for molecular imaging with radionuclides on small animal models has been designed for this aim. A prototype has been built using tungsten pinhole and LaBr3(Ce) scintillator coupled to Hamamatsu Flat Panel PMTs. Compact individual-channel readout has been designed, built and tested. Measurements with phantoms as well as pilot studies on mice have been performed, the results show that the myocardial perfusion in mice can be determined with sufficient precision. The detector will be improved replacing the Hamamatsu Flat Panel with Silicon Photomultipliers (SiPMs) to allow integration of the system with MRI scanners. Application of LaBr3(Ce) scintillator coupled to photosensor with high photon detection efficiency and excellent energy resolution will allow dual-label imaging to monitor simultaneously the cardiac perfusion and the molecular targets under investigation during the heart therapy.

  17. 77 FR 27438 - Certain Corrosion-Resistant Carbon Steel Flat Products From Korea: Final Results of Expedited...

    Science.gov (United States)

    2012-05-10

    ... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Certain Corrosion-Resistant... order on certain corrosion-resistant carbon steel flat products (``CORE'') from the Republic of Korea.... Scope of the Order The merchandise covered by the order includes flat-rolled carbon steel products, of...

  18. Field demonstrations of passive detectors for screening of alpha contaminated soils

    International Nuclear Information System (INIS)

    Meyer, K.E.; Gammage, R.B.; Dudney, C.S.

    1994-01-01

    There are numerous sites around the country, DOE and otherwise, that are faced with the daunting task of remediating radiologically contaminated soils and groundwaters. Some of these sites, such as the Nevada Test Site and the Rocky Flats Plant, have contaminants that have been dispersed over wide areas. The costs of the characterization phase alone for such remediation programs can be prohibitive. Therefore there are pressing needs for testing and evaluation of new technologies for screening for radiological contaminants that may offer significant advantages in capital costs, ease of use, sensitivity, ruggedness, and/or reliability. This work reports on laboratory and Field tests of two types of passive alpha detectors, electret ionization chambers (EIC's) and alpha track detectors (ATD's), that have been commercially developed for indoor radon measurements. Previous work documented calibration and measurement protocols developed for these detectors for indoor surface contamination measurements

  19. Long-range Rocky Flats utilization study

    International Nuclear Information System (INIS)

    1983-02-01

    The purpose of this Study was to provide information concerning the Rocky Flats Plant and its operations that will be useful to the Nation's decision-makers in determining the long-range future of the Plant. This Study was conducted under the premise that national defense policy must be supported and, accordingly, the capabilities at Rocky Flats must be maintained there or at some other location(s). The Study, therefore, makes no attempt to speculate on how possible future changes in national defense policy might affect decisions regarding the utilization of Rocky Flats. Factors pertinent to decisions regarding Rocky Flats, which are included in the Study, are: physical condition of the Plant and its vulnerabilities to natural phenomena; risks associated with plutonium to Plant workers and the public posed by postulated natural phenomena and operational accidents; identification of alternative actions regarding the future use of the Rocky Flats Plant with associated costs and time scales; local socioeconomic impacts if Rocky Flats operations were relocated; and potential for other uses if Rocky Flats facilities were vacated. The results of the tasks performed in support of this Study are summarized in the context of these five factors

  20. Fabrication, characterization and simulation of 4H-SiC Schottky diode alpha particle detectors for pyroprocessing actinide monitoring

    Science.gov (United States)

    Garcia, Timothy Richard

    Pyroprocessing is a method of using high-temperature molten salts and electric fields to separate and collect fuel isotopes of used nuclear fuel. It has been has been tested in the U.S. at Idaho National Laboratory as a key step in closing the nuclear fuel cycle. One technical problem with the pyroprocessing method is a lack of knowledge regarding the actinide concentrations in the salt bath during operation, since on-line techniques for measuring these concentrations are not presently available. 4H-SiC Schottky diode detectors can potentially fulfill this need. Such detectors would operate in contact with the molten salt, and measure concentrations via alpha-particle spectroscopy. This work seeks to fabricate and characterize 4H-SiC Schottky diode detectors at high temperature, model the alpha particle spectrum expected in a molten salt, and model the operation of the detectors to confirm the physics of operation is as expected. In this work, 4H-SiC Schottky diode detectors were fabricated at OSU Nanotech West. After fabrication, these detectors were characterized using both I-V curves and Am-241 alpha-particle energy spectra. All measurements were made as a function of temperature, from room temperature up to 500°C. The average energy required to create an electron-hole pair was observed to decrease with an increase of temperature, due to a decrease of both the 4H-SiC bandgap and non-linear energy loss terms. Furthermore, the FWHM of the spectra was observed to be dependent on the leakage current at a certain temperature, and not dependent on the temperature itself. Secondly, the alpha particle energy spectrum in the pyroprocessing environment was modeled using SRIM. The molten salt was modeled in 3 different geometries, with or without a protective cover material on top of the detector. Due to the loss of alpha-particle energy in the molten salt itself, a high-energy alpha emitter may completely cover the spectrum from a lower-energy alpha emitter. Each of the

  1. Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities.

    Science.gov (United States)

    Cao, Qian; Sisniega, Alejandro; Brehler, Michael; Stayman, J Webster; Yorkston, John; Siewerdsen, Jeffrey H; Zbijewski, Wojciech

    2018-01-01

    Quantitative assessment of trabecular bone microarchitecture in extremity cone-beam CT (CBCT) would benefit from the high spatial resolution, low electronic noise, and fast scan time provided by complementary metal-oxide semiconductor (CMOS) x-ray detectors. We investigate the performance of CMOS sensors in extremity CBCT, in particular with respect to potential advantages of thin (CMOS x-ray detector incorporating the effects of CsI:Tl scintillator thickness was developed. Simulation studies were performed using nominal extremity CBCT acquisition protocols (90 kVp, 0.126 mAs/projection). A range of scintillator thickness (0.35-0.75 mm), pixel size (0.05-0.4 mm), focal spot size (0.05-0.7 mm), magnification (1.1-2.1), and dose (15-40 mGy) was considered. The detectability index was evaluated for both CMOS and a-Si:H flat-panel detector (FPD) configurations for a range of imaging tasks emphasizing spatial frequencies associated with feature size aobj. Experimental validation was performed on a CBCT test bench in the geometry of a compact orthopedic CBCT system (SAD = 43.1 cm, SDD = 56.0 cm, matching that of the Carestream OnSight 3D system). The test-bench studies involved a 0.3 mm focal spot x-ray source and two CMOS detectors (Dalsa Xineos-3030HR, 0.099 mm pixel pitch) - one with the standard CsI:Tl thickness of 0.7 mm (C700) and one with a custom 0.4 mm thick scintillator (C400). Measurements of modulation transfer function (MTF), detective quantum efficiency (DQE), and CBCT scans of a cadaveric knee (15 mGy) were obtained for each detector. Optimal detectability for high-frequency tasks (feature size of ~0.06 mm, consistent with the size of trabeculae) was ~4× for the C700 CMOS detector compared to the a-Si:H FPD at nominal system geometry of extremity CBCT. This is due to ~5× lower electronic noise of a CMOS sensor, which enables input quantum-limited imaging at smaller pixel size. Optimal pixel size for high-frequency tasks was CMOS

  2. Evolution of robotic arms.

    Science.gov (United States)

    Moran, Michael E

    2007-01-01

    The foundation of surgical robotics is in the development of the robotic arm. This is a thorough review of the literature on the nature and development of this device with emphasis on surgical applications. We have reviewed the published literature and classified robotic arms by their application: show, industrial application, medical application, etc. There is a definite trend in the manufacture of robotic arms toward more dextrous devices, more degrees-of-freedom, and capabilities beyond the human arm. da Vinci designed the first sophisticated robotic arm in 1495 with four degrees-of-freedom and an analog on-board controller supplying power and programmability. von Kemplen's chess-playing automaton left arm was quite sophisticated. Unimate introduced the first industrial robotic arm in 1961, it has subsequently evolved into the PUMA arm. In 1963 the Rancho arm was designed; Minsky's Tentacle arm appeared in 1968, Scheinman's Stanford arm in 1969, and MIT's Silver arm in 1974. Aird became the first cyborg human with a robotic arm in 1993. In 2000 Miguel Nicolalis redefined possible man-machine capacity in his work on cerebral implantation in owl-monkeys directly interfacing with robotic arms both locally and at a distance. The robotic arm is the end-effector of robotic systems and currently is the hallmark feature of the da Vinci Surgical System making its entrance into surgical application. But, despite the potential advantages of this computer-controlled master-slave system, robotic arms have definite limitations. Ongoing work in robotics has many potential solutions to the drawbacks of current robotic surgical systems.

  3. Baby-MIND neutrino detector

    Science.gov (United States)

    Mefodiev, A. V.; Kudenko, Yu. G.; Mineev, O. V.; Khotjantsev, A. N.

    2017-11-01

    The main objective of the Baby-MIND detector (Magnetized Iron Neutrino Detector) is the study of muon charge identification efficiency for muon momenta from 0.3 to 5 GeV/ c. This paper presents the results of measurement of the Baby-MIND parameters.

  4. Detection of small pulmonary nodules on chest radiographs: efficacy of dual-energy subtraction technique using flat-panel detector chest radiography

    International Nuclear Information System (INIS)

    Oda, S.; Awai, K.; Funama, Y.; Utsunomiya, D.; Yanaga, Y.; Kawanaka, K.; Nakaura, T.; Hirai, T.; Murakami, R.; Nomori, H.; Yamashita, Y.

    2010-01-01

    Aim: To investigate the effect of a double-exposure dual-energy subtraction (DES) technique on the diagnostic performance of radiologists detecting small pulmonary nodules on flat-panel detector (FPD) chest radiographs. Materials and methods: Using FPD radiography 41 sets of chest radiographs were obtained from 26 patients with pulmonary nodules measuring ≤20 mm and from 15 normal participants. Each dataset included standard and corresponding DES images. There were six non-solid, 10 part-solid, and 10 solid nodules. The mean size of the 26 nodules was 15 ± 4.8 mm. Receiver operating characteristic (ROC) analysis was performed to compare the performance of the eight board-certified radiologists. Results: For the eight radiologists, the mean value of the area under the ROC curve (AUC) without and with DES images was 0.62 ± 0.05 and 0.68 ± 0.05, respectively; the difference was statistically significant (p = 0.02). For part-solid nodules, the difference of the mean AUC value was statistically significant (AUC = 0.61 ± 0.07 versus 0.69 ± 0.05; p < 0.01); for non-solid nodules it was not (AUC = 0.62 ± 0.1 versus 0.61 ± 0.09; p = 0.73), and for solid nodules it was not (AUC = 0.75 ± 0.1 versus 0.78 ± 0.08; p = 0.23). For nodules with overlapping bone shadows, the difference of the mean AUC value was statistically significant (p = 0.03), for nodules without overlapping, it was not (p = 0.26). Conclusion: Use of a double-exposure DES technique at FPD chest radiography significantly improved the diagnostic performance of radiologists to detect small pulmonary nodules.

  5. A Broadband Metasurface-Based Terahertz Flat-Lens Array

    KAUST Repository

    Wang, Qiu; Zhang, Xueqian; Xu, Yuehong; Tian, Zhen; Gu, Jianqiang; Yue, Weisheng; Zhang, Shuang; Han, Jiaguang; Zhang, Weili; Zhang, Weili

    2015-01-01

    A metasurface-based terahertz flat-lens array is proposed, comprising C-shaped split-ring resonators exhibiting locally engineerable phase discontinuities. Possessing a high numerical aperture, the planar lens array is flexible, robust, and shows excellent focusing characteristics in a broadband terahertz frequency. It could be an important step towards the development of planar terahertz focusing devices for practical applications.

  6. A Broadband Metasurface-Based Terahertz Flat-Lens Array

    KAUST Repository

    Wang, Qiu

    2015-02-12

    A metasurface-based terahertz flat-lens array is proposed, comprising C-shaped split-ring resonators exhibiting locally engineerable phase discontinuities. Possessing a high numerical aperture, the planar lens array is flexible, robust, and shows excellent focusing characteristics in a broadband terahertz frequency. It could be an important step towards the development of planar terahertz focusing devices for practical applications.

  7. Readout and trigger for the AFP detector at the ATLAS experiment at LHC

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00097773; The ATLAS collaboration; Kocian, Martin; Lopez Paz, Ivan; Avoni, Giulio

    2017-01-01

    The ATLAS Forward Proton is a new detector system in ATLAS that allows study of events with protons scattered at very small angles. The final design assumes four stations at distances of 205 and 217 m from the ATLAS interaction point on both sides of the detector exploiting the Roman Pot technology. In 2016 two stations in one arm were installed; installation of the other two is planned for 2017. This article describes details of the installed hardware, firmware and software leading to the full integration with the ATLAS central trigger and data acquisition systems.

  8. A genetic linkage map of the chromosome 4 short arm

    Energy Technology Data Exchange (ETDEWEB)

    Locke, P.A.; MacDonald, M.E.; Srinidhi, J.; Tanzi, R.E.; Haines, J.L. (Massachusetts General Hospital, Boston (United States)); Gilliam, T.C. (Columbia Univ., New York, NY (United States)); Conneally, P.M. (Indiana Univ. Medical Center, Indianapolis (United States)); Wexler, N.S. (Columbia Univ., New York, NY (United States) Hereditary Disease Foundation, Santa Monica, CA (United States)); Gusella, J.F. (Massachusetts General Hospital, Boston (United States) Harvard Univ., Boston, MA (United States))

    1993-01-01

    The authors have generated an 18-interval contiguous genetic linkage map of human chromosome 4 spanning the entire short arm and proximal long arm. Fifty-seven polymorphisms, representing 42 loci, were analyzed in the Venezuelan reference pedigree. The markers included seven genes (ADRA2C, ALB, GABRB1, GC, HOX7, IDUA, QDPR), one pseudogene (RAF1P1), and 34 anonymous DNA loci. Four loci were represented by microsatellite polymorphisms and one (GC) was expressed as a protein polymorphism. The remainder were genotyped based on restriction fragment length polymorphism. The sex-averaged map covered 123 cM. Significant differences in sex-specific rates of recombination were observed only in the pericentromeric and proximal long arm regions, but these contributed to different overall map lengths of 115 cM in males and 138 cM in females. This map provides 19 reference points along chromosome 4 that will be particularly useful in anchoring and seeding physical mapping studies and in aiding in disease studies. 26 refs., 1 fig., 1 tab.

  9. Uncooled Radiation Hard Large Area SiC X-ray and EUV Detectors and 2D Arrays, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — This project seeks to design, fabricate, characterize and commercialize large area, uncooled and radiative hard 4H-SiC EUV ? soft X-ray detectors capable of ultra...

  10. Arm-in-Arm Response Regulator Dimers Promote Intermolecular Signal Transduction

    Energy Technology Data Exchange (ETDEWEB)

    Baker, Anna W.; Satyshur, Kenneth A.; Morales, Neydis Moreno; Forest, Katrina T. (UW)

    2016-02-01

    >IMPORTANCEBphP histidine kinases and their cognate response regulators comprise widespread red light-sensing two-component systems. Much work on BphPs has focused on structural understanding of light sensing and on enhancing the natural infrared fluorescence of these proteins, rather than on signal transduction or the resultant phenotypes. To begin to address this knowledge gap, we solved the crystal structures of two single-domain response regulators encoded by a region immediately downstream of that encoding BphPs. We observed a previously unknown arm-in-arm dimer linkage. Monomerization via deletion of the C-terminal dimerization motif had an inhibitory effect on net response regulator phosphorylation, underlining the importance of these unusual dimers for signal transduction.

  11. Modeling and simulation of critical parameters of the first chamber of the dimuon arm spectrometer of the Alice experiment; Modelisation et simulation de parametres critiques de la premiere station du spectrometre dimuons d'ALICE

    Energy Technology Data Exchange (ETDEWEB)

    Guez, D

    2003-10-01

    The Alice experiment that is dedicated to the study of ultra-relativistic heavy ion collisions, will take place in the future large hadron collider (LHC) at CERN. The dimuon arm spectrometer of the Alice experiment is devoted to the search of a new signature of the existence of the quark gluon plasma (QGP). The first chapter is dedicated to the physics notions linked to the study of QGP, a few signatures are proposed for the detection of QGP, particularly the signature concerning the production rate of quarkonium. The second chapter deals with particle detection involved in Alice experiment, the dimuon arm spectrometer is a detector dedicated to the track reconstruction of muons issued from the decay of heavy mesons from J/{psi} and {upsilon} families. The third and the fourth chapters present the studies made to integrate a reliable model of the dimuon arm in the global simulation code of Alice (Aliroot). The fifth chapter presents the software TB{sup 2} that has been developed within the framework of this thesis in order to check and control the output data when the detector is tested with a real particle beam. The sixth chapter presents the results of the tests that have been performed with a 7 GeV/c pion beam. These tests have shown that the electronic noise is coherent with the specifications of Alice experiment. A factor 1,8 between the highest and the weakest values of the gain has been measured in the chamber. The detection efficiency of the chamber has been estimated to 99% in the different cases studied. (A.C.)

  12. Implementation of a dosimetric system based on detectors of to AL{sub 2}O{sub 3}:C and its use in radiation therapy; Implementacion de un sistema dosimetrico basdo en detectores de Al{sub 2}O{sub 3}: C y su uso en radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Marquez, P. J. F.; Guzman, C. C. S.; Pico, C. C.; Aymar, A. J. A.

    2013-07-01

    In this work is proposed to implement an alternative dosimetric system, based on OSL technology of crystals of oxide doped aluminum with carbon, to the AL{sub 2}O{sub 3}:C which do not possess the drawbacks of traditional detectors and possess the advantage for re-reading processes, allowing the storage of the detector as a record of delivered adsorbed dose which may be retested if required. (Author)

  13. Bilateral movements increase sustained extensor force in the paretic arm.

    Science.gov (United States)

    Kang, Nyeonju; Cauraugh, James H

    2018-04-01

    Muscle weakness in the extensors poststroke is a common motor impairment. Unfortunately, research is unclear on whether bilateral movements increase extensor force production in the paretic arm. This study investigated sustained force production while stroke individuals maximally extended their wrist and fingers on their paretic arm. Specifically, we determined isometric force production in three conditions: (a) unilateral paretic arm, (b) unilateral nonparetic arm, and (c) bilateral (both arms executing the same movement simultaneously). Seventeen chronic stroke patients produced isometric sustained force by executing wrist and fingers extension in unilateral and bilateral contraction conditions. Mean force, force variability (coefficient of variation), and signal-to-noise ratio were calculated for each contraction condition. Analysis of two-way (Arm × Type of Condition: 2 × 2; Paretic or Nonparetic Arm × Unilateral or Bilateral Conditions) within-subjects ANOVAs revealed that the bilateral condition increased sustained force in the paretic arm, but reduced sustained force in the nonparetic arm. Further, although the paretic arm exhibited more force variability and less signal-to-noise ratio than the nonparetic arm during a unilateral condition, there were no differences when participants simultaneously executed isometric contractions with both arms. Our unique findings indicate that bilateral contractions transiently increased extensor force in the paretic arm. Implications for Rehabilitation Bilateral movements increased isometric wrsit extensor force in paretic arms and redcued force in nonparetic arms versus unilateral movements. Both paretic and nonparetic arms produced similar force variability and signal-to-noise ratio during bilateral movements. Increased sustained force in the paretic arm during the bilateral condition indicates that rehabilitation protocols based on bilateral movements may be beneficial for functional recovery.

  14. Incorporation of flat glass in red ceramic

    International Nuclear Information System (INIS)

    Caldas, T.C.C.; Morais, A.S.C.; Pereira, P.S.; Monteiro, S.N.; Vieira, C.M.F.

    2011-01-01

    This work have as objective evaluate the effect of incorporation of up to 10% by weight of powdered flat glass , from civil industry, in red ceramic. The bodies were obtained by uniaxial pressing at 20 MPa and fired at temperatures of 850 ° C and 1050 ° C. The parameters studied were linear firing shrinkage, apparent density, water absorption and flexural rupture stress for the evaluation of the mechanical physical properties. The microstructure was observed by scanning electron microscopy and phase identification was performed by X-ray diffraction. The results showed that the waste changes the microstructure and properties of red ceramics. (author)

  15. Trajectory Optimization for Differential Flat Systems

    OpenAIRE

    Kahina Louadj; Benjamas Panomruttanarug; Alexandre Carlos Brandao Ramos; Felix Mora-Camino

    2016-01-01

    International audience; The purpose of this communication is to investigate the applicability of Variational Calculus to the optimization of the operation of differentially flat systems. After introducingcharacteristic properties of differentially flat systems, the applicability of variational calculus to the optimization of flat output trajectories is displayed. Two illustrative examples are also presented.

  16. Development of mirror coatings for gravitational-wave detectors

    Science.gov (United States)

    Steinlechner, J.

    2018-05-01

    Gravitational waves are detected by measuring length changes between mirrors in the arms of kilometre-long Michelson interferometers. Brownian thermal noise arising from thermal vibrations of the mirrors can limit the sensitivity to distance changes between the mirrors, and, therefore, the ability to measure gravitational-wave signals. Thermal noise arising from the highly reflective mirror coatings will limit the sensitivity both of current detectors (when they reach design performance) and of planned future detectors. Therefore, the development of coatings with low thermal noise, which at the same time meet strict optical requirements, is of great importance. This article gives an overview of the current status of coatings and of the different approaches for coating improvement. This article is part of a discussion meeting issue `The promises of gravitational-wave astronomy'.

  17. Rocky Flats Compliance Program

    International Nuclear Information System (INIS)

    1994-02-01

    The Department of Energy (DOE) established the Office of Technology Development (EM-50) (OTD) as an element of Environmental Restoration and Waste Management (EM) in November 1989. The primary objective of the Office of Technology Development, Rocky Flats Compliance Program (RFCP), is to develop altemative treatment technologies for mixed low-level waste (wastes containing both hazardous and radioactive components) to use in bringing the Rocky Flats Plant (RFP) into compliance with Federal and state regulations and agreements. Approximately 48,000 cubic feet of untreated low-level mixed waste, for which treatment has not been specified, are stored at the RFP. The cleanup of the Rocky Flats site is driven by agreements between DOE, the Environmental Protection Agency (EPA), and the Colorado Department of Health (CDH). Under these agreements, a Comprehensive Treatment and Management Plan (CTMP) was drafted to outline the mechanisms by which RFP will achieve compliance with the regulations and agreements. This document describes DOE's strategy to treat low-level mixed waste to meet Land Disposal Restrictions and sets specific milestones related to the regulatory aspects of technology development. These milestones detail schedules for the development of technologies to treat all of the mixed wastes at the RFP. Under the Federal Facilities Compliance Act (FFCA), the CTMP has been incorporated into Rocky Flats Plant Conceptual Site Treatment Plan (CSTP). The CSTP will become the Rocky Flats Plant site Treatment Plan in 1995 and will supersede the CTMP

  18. 49 CFR 231.6 - Flat cars.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 4 2010-10-01 2010-10-01 false Flat cars. 231.6 Section 231.6 Transportation... TRANSPORTATION RAILROAD SAFETY APPLIANCE STANDARDS § 231.6 Flat cars. (Cars with sides 12 inches or less above the floor may be equipped the same as flat cars.) (a) Hand brakes—(1) Number. Same as specified for...

  19. 75 FR 18153 - Corrosion-Resistant Carbon Steel Flat Products from the Republic of Korea: Extension of Time...

    Science.gov (United States)

    2010-04-09

    ... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon Steel Flat Products from the Republic of Korea: Extension of Time Limit for Preliminary Results of... countervailing duty order on corrosion-resistant carbon steel flat products (CORE) from Korea. See Countervailing...

  20. 77 FR 16810 - Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Extension of Time...

    Science.gov (United States)

    2012-03-22

    ... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Extension of Time Limit for Preliminary Results of... Register the countervailing duty order on corrosion-resistant carbon steel flat products (CORE) from Korea...

  1. 76 FR 20954 - Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Extension of Time...

    Science.gov (United States)

    2011-04-14

    ... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Extension of Time Limit for Preliminary Results of... Register the countervailing duty order on corrosion-resistant carbon steel flat products (CORE) from Korea...

  2. 78 FR 55241 - Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Preliminary Results of...

    Science.gov (United States)

    2013-09-10

    ... merchandise covered by this Order \\2\\ is certain corrosion- resistant carbon steel flat products from Korea... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon... the countervailing duty (CVD) order on corrosion-resistant carbon steel flat products (CORE) from the...

  3. Heel Effect: Dose Mapping And Profiling For Mobile C-Arm Fluoroscopy Unit Toshiba SXT-1000A

    International Nuclear Information System (INIS)

    Husaini Salleh; Mohd Khalid Matori; Muhammad Jamal Md Isa; Mohd Ramli Arshad; Shahrul Azlan Azizan; Mohd Firdaus Abdul Rahman; Md Khairusalih Md Zin

    2014-01-01

    Heel Effect is the well known phenomena in x-ray production. It contributes the effect to image formation and as well as scattered radiation. But there is paucity in the study related to heel effect. This study is for mapping and profiling the dose on the surface of water phantom by using mobile C-arm unit Toshiba SXT-1000A. Based on the result the dose profile is increasing up to about 57 % from anode to cathode bound of the irradiated area. This result and information can be used as a guide to manipulate these phenomena for better image quality and radiation safety for this specific and dedicated fluoroscopy unit. (author)

  4. Exact Riemann solutions of the Ripa model for flat and non-flat bottom topographies

    Science.gov (United States)

    Rehman, Asad; Ali, Ishtiaq; Qamar, Shamsul

    2018-03-01

    This article is concerned with the derivation of exact Riemann solutions for Ripa model considering flat and non-flat bottom topographies. The Ripa model is a system of shallow water equations accounting for horizontal temperature gradients. In the case of non-flat bottom topography, the mass, momentum and energy conservation principles are utilized to relate the left and right states across the step-type bottom topography. The resulting system of algebraic equations is solved iteratively. Different numerical case studies of physical interest are considered. The solutions obtained from developed exact Riemann solvers are compared with the approximate solutions of central upwind scheme.

  5. Reassignment of Drosophila willistoni Genome Scaffolds to Chromosome II Arms.

    Science.gov (United States)

    Garcia, Carolina; Delprat, Alejandra; Ruiz, Alfredo; Valente, Vera L S

    2015-10-04

    Drosophila willistoni is a geographically widespread Neotropical species. The genome of strain Gd-H4-1 from Guadeloupe Island (Caribbean) was sequenced in 2007 as part of the 12 Drosophila Genomes Project. The assembled scaffolds were joined based on conserved linkage and assigned to polytene chromosomes based on a handful of genetic and physical markers. This paucity of markers was particularly striking in the metacentric chromosome II, comprised two similarly sized arms, IIL and IIR, traditionally considered homologous to Muller elements C and B, respectively. In this paper we present the cytological mapping of 22 new gene markers to increase the number of markers mapped by in situ hybridization and to test the assignment of scaffolds to the polytene chromosome II arms. For this purpose, we generated, by polymerase chain reaction amplification, one or two gene probes from each scaffold assigned to the chromosome II arms and mapped these probes to the Gd-H4-1 strain's polytene chromosomes by nonfluorescent in situ hybridization. Our findings show that chromosome arms IIL and IIR correspond to Muller elements B and C, respectively, directly contrasting the current homology assignments in D. willistoni and constituting a major reassignment of the scaffolds to chromosome II arms. Copyright © 2015 Garcia et al.

  6. X-ray detectors based on image sensors

    International Nuclear Information System (INIS)

    Costa, A.P.R.

    1983-01-01

    X-ray detectors based on image sensors are described and a comparison is made between the advantages and the disadvantages of such a kind of detectors with the position sensitive detectors. (L.C.) [pt

  7. Projectivity and flatness over the endomorphism ring of a finitely generated comodule

    International Nuclear Information System (INIS)

    Guedenon, T.

    2006-04-01

    Let k be a commutative ring, A a k-algebra, C an A- coring that is projective as a left A-module, *C the dual ring of C and Λ a right C-comodule that is finitely generated as a left *C-module. We give necessary and sufficient conditions for projectivity and flatness of a module over the endomorphism ring End C (Λ). If C contains a grouplike element, we can replace Λ with A. (author)

  8. An improved Monte-Carlo model of the Varian EPID separating support arm and rear-housing backscatter

    International Nuclear Information System (INIS)

    Monville, M E; Greer, P B; Kuncic, Z

    2014-01-01

    Previous investigators of EPID dosimetric properties have ascribed the backscatter, that contaminates dosimetric EPID images, to its supporting arm. Accordingly, Monte-Carlo (MC) EPID models have approximated the backscatter signal from the layers under the detector and the robotic support arm using either uniform or non-uniform solid water slabs, or through convolutions with back-scatter kernels. The aim of this work is to improve the existent MC models by measuring and modelling the separate backscatter contributions of the robotic arm and the rear plastic housing of the EPID. The EPID plastic housing is non-uniform with a 11.9 cm wide indented section that runs across the cross-plane direction in the superior half of the EPID which is 1.75 cm closer to the EPID sensitive layer than the rest of the housing. The thickness of the plastic housing is 0.5 cm. Experiments were performed with and without the housing present by removing all components of the EPID from the housing. The robotic support arm was not present for these measurements. A MC model of the linear accelerator and the EPID was modified to include the rear-housing indentation and results compared to the measurement. The rear housing was found to contribute a maximum of 3% additional signal. The rear housing contribution to the image is non-uniform in the in-plane direction with 2% asymmetry across the central 20 cm of an image irradiating the entire detector. The MC model was able to reproduce this non-uniform contribution. The EPID rear housing contributes a non-uniform backscatter component to the EPID image, which has not been previously characterized. This has been incorporated into an improved MC model of the EPID.

  9. Fourth workshop on experiments and detectors for a relativistic heavy ion collider

    International Nuclear Information System (INIS)

    Fatyga, M.; Moskowitz, B.

    1990-01-01

    This report contains papers on the following topics: physics at RHIC; flavor flow from quark-gluon plasma; space-time quark-gluon cascade; jets in relativistic heavy ion collisions; parton distributions in hard nuclear collisions; experimental working groups, two-arm electron/photon spectrometer collaboration; total and elastic pp cross sections; a 4π tracking TPC magnetic spectrometer; hadron spectroscopy; efficiency and background simulations for J/ψ detection in the RHIC dimuon experiment; the collision regions beam crossing geometries; Monte Carlo simulations of interactions and detectors; proton-nucleus interactions; the physics of strong electromagnetic fields in collisions of relativistic heavy ions; a real time expert system for experimental high energy/nuclear physics; the development of silicon multiplicity detectors; a pad readout detector for CRID/tracking; RHIC TPC R ampersand D progress and goals; development of analog memories for RHIC detector front-end electronic systems; calorimeter/absorber optimization for a RHIC dimuon experiment; construction of a highly segmented high resolution TOF system; progress report on a fast, particle-identifying trigger based on ring-imaging and highly integrated electronics for a TPC detector

  10. Fourth workshop on experiments and detectors for a relativistic heavy ion collider

    Energy Technology Data Exchange (ETDEWEB)

    Fatyga, M.; Moskowitz, B. (eds.)

    1990-01-01

    This report contains papers on the following topics: physics at RHIC; flavor flow from quark-gluon plasma; space-time quark-gluon cascade; jets in relativistic heavy ion collisions; parton distributions in hard nuclear collisions; experimental working groups, two-arm electron/photon spectrometer collaboration; total and elastic pp cross sections; a 4{pi} tracking TPC magnetic spectrometer; hadron spectroscopy; efficiency and background simulations for J/{psi} detection in the RHIC dimuon experiment; the collision regions beam crossing geometries; Monte Carlo simulations of interactions and detectors; proton-nucleus interactions; the physics of strong electromagnetic fields in collisions of relativistic heavy ions; a real time expert system for experimental high energy/nuclear physics; the development of silicon multiplicity detectors; a pad readout detector for CRID/tracking; RHIC TPC R D progress and goals; development of analog memories for RHIC detector front-end electronic systems; calorimeter/absorber optimization for a RHIC dimuon experiment; construction of a highly segmented high resolution TOF system; progress report on a fast, particle-identifying trigger based on ring-imaging and highly integrated electronics for a TPC detector.

  11. 78 FR 19210 - Corrosion-Resistant Carbon Steel Flat Products From the Republic of Korea: Final Results of...

    Science.gov (United States)

    2013-03-29

    .... Scope of the Order Products covered by this order are certain corrosion-resistant carbon steel flat... DEPARTMENT OF COMMERCE International Trade Administration [C-580-818] Corrosion-Resistant Carbon... countervailing duty (CVD) order on corrosion-resistant carbon steel flat products from the Republic of Korea for...

  12. Detectors - Electronics; Detecteurs - Electronique

    Energy Technology Data Exchange (ETDEWEB)

    Bregeault, J.; Gabriel, J.L.; Hierle, G.; Lebotlan, P.; Leconte, A.; Lelandais, J.; Mosrin, P.; Munsch, P.; Saur, H.; Tillier, J. [Lab. de Physique Corpusculaire, Caen Univ., 14 (France)

    1998-04-01

    The reports presents the main results obtained in the fields of radiation detectors and associated electronics. In the domain of X-ray gas detectors for the keV range efforts were undertaken to rise the detector efficiency. Multiple gap parallel plate chambers of different types as well as different types of X {yields} e{sup -} converters were tested to improve the efficiency (values of 2.4% at 60 KeV were reached). In the field of scintillators a study of new crystals has been carried out (among which Lutetium orthosilicate). CdTe diode strips for obtaining X-ray imaging were studied. The complete study of a linear array of 8 CdTe pixels has been performed and certified. The results are encouraging and point to this method as a satisfying solution. Also, a large dimension programmable chamber was used to study the influence of temperature on the inorganic scintillators in an interval from -40 deg. C to +150 deg. C. Temperature effects on other detectors and electronic circuits were also investigated. In the report mentioned is also the work carried out for the realization of the DEMON neutron multidetector. For neutron halo experiments different large area Si detectors associated with solid and gas position detectors were realized. In the frame of a contract with COGEMA a systematic study of Li doped glasses was undertaken aiming at replacing with a neutron probe the {sup 3}He counters presently utilized in pollution monitoring. An industrial prototype has been realised. Other studies were related to integrated analog chains, materials for Cherenkov detectors, scintillation probes for experiments on fundamental processes, gas position sensitive detectors, etc. In the field of associated electronics there are mentioned the works related to the multidetector INDRA, data acquisition, software gamma spectrometry, automatic gas pressure regulation in detectors, etc

  13. A new experiment-agnostic mechanism to persistify and serve the detector geometry of ATLAS

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00211497; The ATLAS collaboration; Boudreau, Joseph; Vukotic, Ilija

    2017-01-01

    The complex geometry of the whole detector of the ATLAS experiment at LHC is currently stored only in custom online databases, from which it is built on-the-fly on request. Accessing the online geometry guarantees accessing the latest version of the detector description, but requires the setup of the full ATLAS software framework "Athena", which provides the online services and the tools to retrieve the data from the database. This operation is cumbersome and slows down the applications that need to access the geometry. Moreover, all applications that need to access the detector geometry need to be built and run on the same platform as the ATLAS framework, preventing the usage of the actual detector geometry in stand-alone applications. Here we propose a new mechanism to persistify and serve the geometry of HEP experiments. The new mechanism is composed by a new file format and the modules to make use of it. The new file format allows to store the whole detector description locally in a flat file, and it is e...

  14. Fast and effective embedded systems design applying the ARM mbed

    CERN Document Server

    Toulson, Rob

    2012-01-01

    A hands-on introduction to the field of embedded systems; A focus on fast prototyping of embedded systems; All key embedded system concepts covered through simple and effective experimentation; An understanding of ARM technology, one of the world's leaders; A practical introduction to embedded C; Applies possibly the most accessible set of tools available in the embedded world.  This book is an introduction to embedded systems design, using the ARM mbed and C programming language as development tools. The mbed provides a compact, self-contained and low-cost hardware core, and the

  15. Increasing the compressive strength of portland cement concrete using flat glass powder

    Energy Technology Data Exchange (ETDEWEB)

    Miranda Junior, Edson Jansen Pedrosa de; Bezerra, Helton de Jesus Costa Leite; Politi, Flavio Salgado; Paiva, Antonio Ernandes Macedo, E-mail: edson.jansen@ifma.edu.br [Instituto Federal de Educacao, Ciencia e Tecnologia do Maranha (IFMA), Sao Luis, MA (Brazil). Dept. de Mecanica e Materiais

    2014-08-15

    This paper analyzes the compressive strength of Portland cement concrete in response to the incorporation of 5%, 10% and 20% of flat glass powder in place of sand, at w/c (water/cement) ratios of 0.50, 0.55 and 0.58. A statistical analysis of variance (ANOVA) was performed after 7, 14 and 28 days of curing. The compressive strength test results indicate that the concrete containing a w/c ratio of 0.50 can be used for structural applications, regardless of the waste glass content, as can that with a w/c ratio of 0.55 containing 20% of waste glass. We suggest that the use of flat glass powder in place of sand in the above mentioned percentages is feasible for the production of an environmentally appropriate and structurally applicable concrete. However, the concrete's fluidity and void content must be taken into account. (author)

  16. Have Third-World Arms Industries Reduced Arms Imports?

    OpenAIRE

    Looney, R.E.

    1989-01-01

    Current Research on Peace and Violence, no. 1, 1989. Refereed Journal Article In 1945 only Argentina, Brazil, India and South Africa in the Third World possessed domestic arms industries which produced weapons systems other than small arms and ammunition (SIPRI, 1987, 76).

  17. Design and spectrum calculation of 4H-SiC thermal neutron detectors using FLUKA and TCAD

    Science.gov (United States)

    Huang, Haili; Tang, Xiaoyan; Guo, Hui; Zhang, Yimen; Zhang, Yimeng; Zhang, Yuming

    2016-10-01

    SiC is a promising material for neutron detection in a harsh environment due to its wide band gap, high displacement threshold energy and high thermal conductivity. To increase the detection efficiency of SiC, a converter such as 6LiF or 10B is introduced. In this paper, pulse-height spectra of a PIN diode with a 6LiF conversion layer exposed to thermal neutrons (0.026 eV) are calculated using TCAD and Monte Carlo simulations. First, the conversion efficiency of a thermal neutron with respect to the thickness of 6LiF was calculated by using a FLUKA code, and a maximal efficiency of approximately 5% was achieved. Next, the energy distributions of both 3H and α induced by the 6LiF reaction according to different ranges of emission angle are analyzed. Subsequently, transient pulses generated by the bombardment of single 3H or α-particles are calculated. Finally, pulse height spectra are obtained with a detector efficiency of 4.53%. Comparisons of the simulated result with the experimental data are also presented, and the calculated spectrum shows an acceptable similarity to the experimental data. This work would be useful for radiation-sensing applications, especially for SiC detector design.

  18. Laboratory Measurement of the Brighter-fatter Effect in an H2RG Infrared Detector

    Science.gov (United States)

    Plazas, A. A.; Shapiro, C.; Smith, R.; Huff, E.; Rhodes, J.

    2018-06-01

    The “brighter-fatter” (BF) effect is a phenomenon—originally discovered in charge coupled devices—in which the size of the detector point-spread function (PSF) increases with brightness. We present, for the first time, laboratory measurements demonstrating the existence of the effect in a Hawaii-2RG HgCdTe near-infrared (NIR) detector. We use JPL’s Precision Projector Laboratory, a facility for emulating astronomical observations with UV/VIS/NIR detectors, to project about 17,000 point sources onto the detector to stimulate the effect. After calibrating the detector for nonlinearity with flat-fields, we find evidence that charge is nonlinearly shifted from bright pixels to neighboring pixels during exposures of point sources, consistent with the existence of a BF-type effect. NASAs Wide Field Infrared Survey Telescope (WFIRST) will use similar detectors to measure weak gravitational lensing from the shapes of hundreds of million of galaxies in the NIR. The WFIRST PSF size must be calibrated to ≈0.1% to avoid biased inferences of dark matter and dark energy parameters; therefore further study and calibration of the BF effect in realistic images will be crucial.

  19. F.P.G.: Fastbus Pattern Generator and Bus Configuration detector

    International Nuclear Information System (INIS)

    Cerrito, L.; Chorowicz, V.; Lebbolo, H.; Tesseidre, A.

    1986-03-01

    This module has been developed for the DELPHI-OUTER DETECTOR read out system environment. It is a Fastbus slave designed as a tool for syncronizing and monitoring a complex Fastbus system. F.P.G. is a programmable pattern generator with front panel TTL output: six on lemo cable and 24 on flat cable. Its frequency can range between 67 KHz and 20 MHz. It can be programmed to generate a signal on occurency of a particular configuration of the crate bus

  20. Particle detectors based on semiconducting InP epitaxial layers

    Czech Academy of Sciences Publication Activity Database

    Yatskiv, Roman; Grym, Jan; Žďánský, Karel

    2011-01-01

    Roč. 6, C01072 (2011), C010721-C010725 ISSN 1748-0221 R&D Projects: GA AV ČR KJB200670901; GA MŠk(CZ) OC10021; GA ČR(CZ) GP102/08/P617 Institutional research plan: CEZ:AV0Z20670512 Keywords : Solid state detectors * Gamma detectors * Radiation-hard detectors Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering Impact factor: 1.869, year: 2011