WorldWideScience

Sample records for borne x-ray detectors

  1. A balloon-borne solid state cosmic X-ray detector

    International Nuclear Information System (INIS)

    Proctor, R.; Pietsch, W.; Reppin, C.

    1982-01-01

    On 9th May 1980 a MPI/AIT hard X-ray balloon payload successfully observed numerous cosmic X-ray sources. The payload consisted of a 2400 cm 2 Phoswich detector and a 114 cm 2 solid state detector. The solid state detector is described in this report. It consists of six intrinsic germanium planar crystals in a vacuum cryostat cooled by liquid nitrogen. The detector operates in the hard X-ray energy range of 20-150 keV and had in-flight a mean energy resolution of 2.75 keV at 60 keV. A hexagonal molybdenum collimator defined the field of view as approximately 4 0 fwhm. A CsI(Na) and plastic active shield and passive shielding provided background rejection. Mean background values of 1.3 X 10 -3 counts/(sec x cm 2 x keV) at 60 keV were obtained. (orig.)

  2. Simultaneous investigation of parent electrons and bremsstrahlung x rays by rocket-borne detectors

    International Nuclear Information System (INIS)

    Vij, K.K.; Venkatesan, D.; Sheldon, W.R.; Kern, J.W.; Benbrook, J.R.; Whalen, B.A.

    1975-01-01

    Simultaneous measurements of the energy spectrum of precipitating electrons and the resulting bremsstrahlung X ray spectrum were carried out during an auroral event on March 3, 1971, at the Churchill Research Range, Manitoba, Canada. The electron data were obtained with detectors on a Black Brant VB sounding rocket (275-km apogee), while the X ray flux was measured by an instrument package that was boosted to 60 km on an Arcas rocket. The X ray package was deployed on a parachute at apogee to provide a slow descent through the atmosphere. Thick target bremsstrahlung theory is used to calculate the X ray flux produced by the incident electrons, and a Monte Carlo method is used to predict the X ray spectrum at various altitudes appropriate for comparison with the measured X ray data. Satisfactory agreement between theory and experiment is obtained, and the value of the constant in the thick target theory has been estimated to be (2plus-or-minus0.5) times10 -5

  3. Simultaneous investigation of parent electrons and bremsstrahlung x rays by Rocket--Borne detectors

    International Nuclear Information System (INIS)

    Vij, K.K.; Venkatesan, D.; Sheldon, W.R.; Kern, J.W.; Benbrook, J.R.; Whalen, B.A.

    1975-01-01

    Simultaneous measurements of the energy spectrum of precipitating electrons and the resulting bremsstrahlung X ray spectrum were carried out during an auroral event on March 3, 1971, at the Churchill Research Range, Manitoba, Canada. The electron data were obtained with detectors on a Black Brant VB sounding rocket (275-km apogee), while the X ray flux was measured by an instrument package that was boosted to 60 km on an Arcas rocket. The X ray package was deployed on a parachute apogee to provide a slow descent through the atmosphere. Thick target bremsstrahlung theory is used to calculate the X ray flux produced by the incident electrons, and a Monte Carlo method is used to predict the X ray spectrum at various altitudes appropriate for comparison with the measured X ray data. Satisfactory agreement between theory and experiment is obtained, and the value of the constant in the thick target theory has been estimated to be (2plus-or-minus0.5) times10 -5 . (auth)

  4. X-ray detector array

    International Nuclear Information System (INIS)

    Houston, J.M.

    1980-01-01

    The object of the invention (an ionization chamber X-ray detector array for use with high speed computerised tomographic imaging apparatus) is to reduce the time required to produce a tomographic image. The detector array described determines the distribution of X-ray intensities in one or more flat, coplanar X-ray beams. It comprises three flat anode sheets parallel to the X-ray beam, a plurality of rod-like cathodes between the anodes, a detector gas between the electrodes and a means for applying a potential between the electrodes. Each of the X-ray sources is collimated to give a narrow, planar section of X-ray photons. Sets of X-ray sources in the array are pulsed simultaneously to obtain X-ray transmission data for tomographic image reconstruction. (U.K.)

  5. X-ray detector

    International Nuclear Information System (INIS)

    Whetten, N.R.; Houston, J.M.

    1977-01-01

    An ionization chamber for use in determining the spatial distribution of x-ray photons in tomography systems comprises a plurality of substantially parallel, planar anodes separated by parallel, planar cathodes and enclosed in a gas of high atomic weight at a pressure from approximately 10 atmospheres to approximately 50 atmospheres. The cathode and anode structures comprise metals which are substantially opaque to x-ray radiation and thereby tend to reduce the resolution limiting effects of x-ray fluoresence in the gas. In another embodiment of the invention the anodes comprise parallel conductive bars disposed between two planar cathodes. Guard rings eliminate surface leakage currents between adjacent electrodes. 8 figures

  6. X-ray detector

    International Nuclear Information System (INIS)

    Houston, J.M.; Whetten, N.R.

    1981-01-01

    An ionization chamber for use in determining the spatial distribution of x-ray photons in tomography systems comprises a plurality of substantially parallel, planar anodes separated by parallel, planar cathodes and enclosed in a gas of high atomic weight at a pressure from approximately 10 atmospheres to approximately 50 atmospheres. The cathode and anode structures comprise metals which are substantially opaque to x-ray radiation and thereby tend to reduce the resolution limiting effects of xray fluoresence in the gas. In another embodiment of the invention the anodes comprise parallel conductive bars disposed between two planar cathodes. Guard rings eliminate surface leakage currents between adjacent electrodes

  7. Semiconductor X-ray detectors

    CERN Document Server

    Lowe, Barrie Glyn

    2014-01-01

    Identifying and measuring the elemental x-rays released when materials are examined with particles (electrons, protons, alpha particles, etc.) or photons (x-rays and gamma rays) is still considered to be the primary analytical technique for routine and non-destructive materials analysis. The Lithium Drifted Silicon (Si(Li)) X-Ray Detector, with its good resolution and peak to background, pioneered this type of analysis on electron microscopes, x-ray fluorescence instruments, and radioactive source- and accelerator-based excitation systems. Although rapid progress in Silicon Drift Detectors (SDDs), Charge Coupled Devices (CCDs), and Compound Semiconductor Detectors, including renewed interest in alternative materials such as CdZnTe and diamond, has made the Si(Li) X-Ray Detector nearly obsolete, the device serves as a useful benchmark and still is used in special instances where its large, sensitive depth is essential. Semiconductor X-Ray Detectors focuses on the history and development of Si(Li) X-Ray Detect...

  8. Investigation of pulse shape analyzers for phoswich detectors in space-borne hard X-ray experiments

    Energy Technology Data Exchange (ETDEWEB)

    Bleeker, J A.M.; Overtoom, J M [Huygens Lab., Leiden (Netherlands). Cosmic Ray Working Group

    1979-12-01

    A low-background telescope for hard X-ray astronomy (15-250 keV), comprising arrays of NaI(Tl)/CsI(Na) phoswiches as photon collectors, was recently developed. The background rejection efficiency of such a telescope, and hence the minimum source in a given time, critically depends on the performance of the phoswich pulse shape analyzer (PSA) in a space radiation environment. Results from theoretical and experimental work on analyzer configurations based on zero-crossing detection are presented. This led to the selection of an optimum configuration for space application. The in-situ performance of this analyzer was evaluated in a balloon-borne hard X-ray experiment, showing excellent discrimination efficiency throughout the entire energy regime.

  9. Si(Li) X-ray detector

    International Nuclear Information System (INIS)

    Yuan Xianglin; Li Zhiyong; Hong Xiuse

    1990-08-01

    The fabrication technology of the 10∼80 mm 2 Si(Li) X-ray detectors are described and some problems concerning technology and measurement are discussed. The specifications of the detectors are shown as well. The Si(Li) X-ray detector is a kind of low energy X-ray detectors. Owing to very high energy resolution, fine linearity and high detection efficiency in the range of low energy X-rays, it is widely used in the fields of nuclear physics, medicine, geology and environmental protection, etc,. It is also a kernel component for the scanning electron microscope and X-ray fluorescence analysis systems

  10. Gas position sensitive x-ray detectors

    International Nuclear Information System (INIS)

    Barbosa, A.F.

    1994-12-01

    The construction of gas x-ray detectors used to count and localize x-ray photons in one and two dimensions is reported. The principles of operation of the detectors are described, as well as the electronic modules comprised in the data acquisition system. Results obtained with detectors built at CBPF are shown, illustrating the performance of the Linear Position Sensitive Detectors. (author). 6 refs, 14 figs

  11. Multichannel X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Khabakhpashev, A

    1980-10-01

    A typical design is discussed of multiwire proportional counters and their characteristic feature is explained, ie., the possibility of showing one or two coordinates of the X-ray quantum absorption site. The advantages of such instruments are listed, such as increased sensitivity of determination, the possibility of recording radiations of a different intensity, the possibility of on-line data processing and of the digital display of results. The fields of application include X-ray structural analysis in solid state physics, crystallography, molecular biology, astronomy, materials testing, and medicine.

  12. Detector unit for X-ray diagnosis

    International Nuclear Information System (INIS)

    Svobodova, B.; Hamouz, J.; Pavlicek, Z.; Jursova, L.; Pohanka, J.

    1983-01-01

    The detector unit is applied in the medical and industrial X-ray diagnosis and analysis. It controls the X-ray dosing by exposure and brightness automation. The detector field is generated from a carrier, in which detector elements with light quides are situated, tapped on optical detectors with level converters outside the detector field. The detector field and the optical detectors with level converters are located in a light-resistent shell. This arrangement of the detector unit allows to use the impulse skiascopy instead of permanent X-ray examinations or the skiagraphy with multienergy levels which considerably improves the diagnostic value of the exposures and the working conditions. 1 cl., 1 fig

  13. Improvements in X-ray detectors

    International Nuclear Information System (INIS)

    Whetten, N.R.; Houston, J.M.

    1979-01-01

    Multicellular, spatially separate, gaseous ionization detectors for use in computerized tomography are described. They have high sensitivity, short recovery time, fine spatial resolution and are relatively insensitive to the adverse effects of k shell x-ray fluoresecence.(UK)

  14. Position-sensitive X-ray detectors

    International Nuclear Information System (INIS)

    Hendrix, J.

    1982-01-01

    An overview is given of the different types of position-sensitive X-ray detectors used in kinetic studies of biological molecule state changes using X-ray diffraction with synchrotron radiation as a probe. The detector requirements and principles of operation of proportional counters are outlined. Multiwire proportional chamber systems and their readout techniques are described. Other detectors discussed include a drift chamber type detector, microchannel plates, charge-couple devices and, for high count rates, an integrating TV-detector. (U.K.)

  15. Scintillating ribbon x-ray detector

    International Nuclear Information System (INIS)

    Kinchen, B.E.; Rogers, A.

    1995-01-01

    A patent in the early 1970's by Aerojet Corporation in Sacramento, CA put forth the idea of using an array of scintillating fibers for x-ray detection and imaging. In about 1975, Pratt and Whitney Aircraft in East Hartford, CT designed and manufactured an imaging system based on the patent. The device was 1.75 in thick in the direction of the x-ray beam and about 4 in. by 4 in. square. The device was used with a 8 MeV x-ray source to image and measure internal clearances within operating aircraft, gas turbines engines. There are significant advantages of fiber optic detectors in x-ray detection. However, the advantages are often outweighed by the disadvantages. Two of the advantages of scintillating fiber optic x-ray detectors are: (1) high limiting spatial frequency -- between 20 and 25 lp/mm; and (2) excellent x-ray stopping power -- they can be made thick and retain spatial resolution. In traditional fiber optic detectors the x-rays are oriented parallel to the long axis of the fiber. For the scintillating ribbon x-ray sensor, the x-rays are oriented normal to the fiber long axis. This ribbon sensor technique has a number of advantages over the two current radiographic techniques digital x-radiography and x-ray film: The main advantage the ribbon has is size and shape. It can be as thin as 0.05 in., virtually any width or length, and flexible. Once positioned in a given location, 20 to 100 square inches of the object being inspected can be imaged with a single x-ray beam sweep. It is clear that conventional digital cameras do not lend themselves to placement between walls of aircraft structures or similar items requiring x-ray inspections. A prototype scintillating ribbon x-ray sensor has been fabricated and tested by Synergistic Detector Designs. Images were acquired on corrosion test panels of aluminum fabricated by Iowa State University

  16. Multicell x-ray detector

    International Nuclear Information System (INIS)

    Stone, B.N.; Shelley, P.S.; Love, W.D.

    1981-01-01

    This invention is concerned with improving multicell detectors, particularly those used in computerized tomography. Existing ionization detectors have problems maintaining the precise dimensional spacing between electrodes required for accuracy. In addition, mechanical vibrations set up microphonic effects between the electrode plates. In this invention, pairs of electrode plates are separated by grooved insulating members. The upper and lower edges of an array of electrode plates are inserted in corresponding grooves in the insulating members, and, the whole electrode assembly is securely anchored in the detector chamber

  17. Pixel detectors for x-ray imaging spectroscopy in space

    International Nuclear Information System (INIS)

    Treis, J; Andritschke, R; Hartmann, R; Herrmann, S; Holl, P; Lauf, T; Lechner, P; Lutz, G; Meidinger, N; Porro, M; Richter, R H; Schopper, F; Soltau, H; Strueder, L

    2009-01-01

    Pixelated semiconductor detectors for X-ray imaging spectroscopy are foreseen as key components of the payload of various future space missions exploring the x-ray sky. Located on the platform of the new Spectrum-Roentgen-Gamma satellite, the eROSITA (extended Roentgen Survey with an Imaging Telescope Array) instrument will perform an imaging all-sky survey up to an X-ray energy of 10 keV with unprecedented spectral and angular resolution. The instrument will consist of seven parallel oriented mirror modules each having its own pnCCD camera in the focus. The satellite born X-ray observatory SIMBOL-X will be the first mission to use formation-flying techniques to implement an X-ray telescope with an unprecedented focal length of around 20 m. The detector instrumentation consists of separate high- and low energy detectors, a monolithic 128 x 128 DEPFET macropixel array and a pixellated CdZTe detector respectively, making energy band between 0.5 to 80 keV accessible. A similar concept is proposed for the next generation X-ray observatory IXO. Finally, the MIXS (Mercury Imaging X-ray Spectrometer) instrument on the European Mercury exploration mission BepiColombo will use DEPFET macropixel arrays together with a small X-ray telescope to perform a spatially resolved planetary XRF analysis of Mercury's crust. Here, the mission concepts and their scientific targets are briefly discussed, and the resulting requirements on the detector devices together with the implementation strategies are shown.

  18. Pixel detectors for x-ray imaging spectroscopy in space

    Science.gov (United States)

    Treis, J.; Andritschke, R.; Hartmann, R.; Herrmann, S.; Holl, P.; Lauf, T.; Lechner, P.; Lutz, G.; Meidinger, N.; Porro, M.; Richter, R. H.; Schopper, F.; Soltau, H.; Strüder, L.

    2009-03-01

    Pixelated semiconductor detectors for X-ray imaging spectroscopy are foreseen as key components of the payload of various future space missions exploring the x-ray sky. Located on the platform of the new Spectrum-Roentgen-Gamma satellite, the eROSITA (extended Roentgen Survey with an Imaging Telescope Array) instrument will perform an imaging all-sky survey up to an X-ray energy of 10 keV with unprecedented spectral and angular resolution. The instrument will consist of seven parallel oriented mirror modules each having its own pnCCD camera in the focus. The satellite born X-ray observatory SIMBOL-X will be the first mission to use formation-flying techniques to implement an X-ray telescope with an unprecedented focal length of around 20 m. The detector instrumentation consists of separate high- and low energy detectors, a monolithic 128 × 128 DEPFET macropixel array and a pixellated CdZTe detector respectively, making energy band between 0.5 to 80 keV accessible. A similar concept is proposed for the next generation X-ray observatory IXO. Finally, the MIXS (Mercury Imaging X-ray Spectrometer) instrument on the European Mercury exploration mission BepiColombo will use DEPFET macropixel arrays together with a small X-ray telescope to perform a spatially resolved planetary XRF analysis of Mercury's crust. Here, the mission concepts and their scientific targets are briefly discussed, and the resulting requirements on the detector devices together with the implementation strategies are shown.

  19. Pixel detectors for x-ray imaging spectroscopy in space

    Energy Technology Data Exchange (ETDEWEB)

    Treis, J; Andritschke, R; Hartmann, R; Herrmann, S; Holl, P; Lauf, T; Lechner, P; Lutz, G; Meidinger, N; Porro, M; Richter, R H; Schopper, F; Soltau, H; Strueder, L [MPI Semiconductor Laboratory, Otto-Hahn-Ring 6, D-81739 Munich (Germany)], E-mail: jft@hll.mpg.de

    2009-03-15

    Pixelated semiconductor detectors for X-ray imaging spectroscopy are foreseen as key components of the payload of various future space missions exploring the x-ray sky. Located on the platform of the new Spectrum-Roentgen-Gamma satellite, the eROSITA (extended Roentgen Survey with an Imaging Telescope Array) instrument will perform an imaging all-sky survey up to an X-ray energy of 10 keV with unprecedented spectral and angular resolution. The instrument will consist of seven parallel oriented mirror modules each having its own pnCCD camera in the focus. The satellite born X-ray observatory SIMBOL-X will be the first mission to use formation-flying techniques to implement an X-ray telescope with an unprecedented focal length of around 20 m. The detector instrumentation consists of separate high- and low energy detectors, a monolithic 128 x 128 DEPFET macropixel array and a pixellated CdZTe detector respectively, making energy band between 0.5 to 80 keV accessible. A similar concept is proposed for the next generation X-ray observatory IXO. Finally, the MIXS (Mercury Imaging X-ray Spectrometer) instrument on the European Mercury exploration mission BepiColombo will use DEPFET macropixel arrays together with a small X-ray telescope to perform a spatially resolved planetary XRF analysis of Mercury's crust. Here, the mission concepts and their scientific targets are briefly discussed, and the resulting requirements on the detector devices together with the implementation strategies are shown.

  20. X-ray detectors in medical imaging

    International Nuclear Information System (INIS)

    Spahn, Martin

    2013-01-01

    Healthcare systems are subject to continuous adaptation, following trends such as the change of demographic structures, the rise of life-style related and chronic diseases, and the need for efficient and outcome-oriented procedures. This also influences the design of new imaging systems as well as their components. The applications of X-ray imaging in the medical field are manifold and have led to dedicated modalities supporting specific imaging requirements, for example in computed tomography (CT), radiography, angiography, surgery or mammography, delivering projection or volumetric imaging data. Depending on the clinical needs, some X-ray systems enable diagnostic imaging while others support interventional procedures. X-ray detector design requirements for the different medical applications can vary strongly with respect to size and shape, spatial resolution, frame rates and X-ray flux, among others. Today, integrating X-ray detectors are in common use. They are predominantly based on scintillators (e.g. CsI or Gd 2 O 2 S) and arrays of photodiodes made from crystalline silicon (Si) or amorphous silicon (a-Si) or they employ semiconductors (e.g. Se) with active a-Si readout matrices. Ongoing and future developments of X-ray detectors will include optimization of current state-of-the-art integrating detectors in terms of performance and cost, will enable the usage of large size CMOS-based detectors, and may facilitate photon counting techniques with the potential to further enhance performance characteristics and foster the prospect of new clinical applications

  1. Neutron and X-ray Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Carini, Gabriella [SLAC National Accelerator Lab., Menlo Park, CA (United States); Denes, Peter [Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States); Gruener, Sol [Cornell Univ., Ithaca, NY (United States); Lessner, Elianne [Dept. of Energy (DOE), Washington DC (United States). Office of Science Office of Basic Energy Sciences

    2012-08-01

    The Basic Energy Sciences (BES) X-ray and neutron user facilities attract more than 12,000 researchers each year to perform cutting-edge science at these state-of-the-art sources. While impressive breakthroughs in X-ray and neutron sources give us the powerful illumination needed to peer into the nano- to mesoscale world, a stumbling block continues to be the distinct lag in detector development, which is slowing progress toward data collection and analysis. Urgently needed detector improvements would reveal chemical composition and bonding in 3-D and in real time, allow researchers to watch “movies” of essential life processes as they happen, and make much more efficient use of every X-ray and neutron produced by the source The immense scientific potential that will come from better detectors has triggered worldwide activity in this area. Europe in particular has made impressive strides, outpacing the United States on several fronts. Maintaining a vital U.S. leadership in this key research endeavor will require targeted investments in detector R&D and infrastructure. To clarify the gap between detector development and source advances, and to identify opportunities to maximize the scientific impact of BES user facilities, a workshop on Neutron and X-ray Detectors was held August 1-3, 2012, in Gaithersburg, Maryland. Participants from universities, national laboratories, and commercial organizations from the United States and around the globe participated in plenary sessions, breakout groups, and joint open-discussion summary sessions. Sources have become immensely more powerful and are now brighter (more particles focused onto the sample per second) and more precise (higher spatial, spectral, and temporal resolution). To fully utilize these source advances, detectors must become faster, more efficient, and more discriminating. In supporting the mission of today’s cutting-edge neutron and X-ray sources, the workshop identified six detector research challenges

  2. Small Angle X-Ray Scattering Detector

    Energy Technology Data Exchange (ETDEWEB)

    Hessler, Jan P.

    2004-06-15

    A detector for time-resolved small-angle x-ray scattering includes a nearly constant diameter, evacuated linear tube having an end plate detector with a first fluorescent screen and concentric rings of first fiber optic bundles for low angle scattering detection and an annular detector having a second fluorescent screen and second fiber optic bundles concentrically disposed about the tube for higher angle scattering detection. With the scattering source, i.e., the specimen under investigation, located outside of the evacuated tube on the tube's longitudinal axis, scattered x-rays are detected by the fiber optic bundles, to each of which is coupled a respective photodetector, to provide a measurement resolution, i.e., dq/q, where q is the momentum transferred from an incident x-ray to an x-ray scattering specimen, of 2% over two (2) orders of magnitude in reciprocal space, i.e., qmax/qmin approx=lO0.

  3. Position sensitive x-ray detector

    International Nuclear Information System (INIS)

    Macchione, E.L.A.

    1990-01-01

    A multi ware position sensitive gas counter for X-ray detection was developed in our laboratory, making use of commercial delay-lines for position sensing. Six delay-line chips (50 ns delay each, 40 Mhz cut-off frequency) cover a total sensitive length of 150 mm leading to a delay-risetime ratio that allows for a high-resolution position detection. Tests using the 5,9 keV X-ray line from a 55 Fe source and integral linearity better than 0,1% and a maximal differential linearity of ±4,0% were obtained operating the detector with an Ar-C H 4 (90%-10%) gas mixture at 700 torr. Similar tests were performed, using the 8,04 keV line from a Cu x-ray tube. A total resolution of 330 μm, and the same integral and differential linearities were obtained. (author)

  4. Silicon lithium detector for x ray fluorescence

    International Nuclear Information System (INIS)

    Rodriguez Cabal, A. E.; Diaz Garcia, A.; Noriega Scull, C.; Martinez Munoz, O.; Diaz Cepeda, R.

    1997-01-01

    The Silicon Lithium detector is the system for the detection of nuclear radiation. It transforms the charge that was produced inside of Silicon material as a result of the incidence of particles and X rays, in voltage pulses at the output of the preamplifier. In this work was made the adjustment of the technological process of manufacture of the detector. Also was made the design and construction of the cryostat and preamplifier and then the validation of the system in a Cuban Dewar. The system, which was made for the first time in our country, has an energy resolution of 185 eV for the Fe-55 source (E=5.9 KeV), which has permitted its implementation in energy dispersive X ray fluorescence. (author) [es

  5. Ultra high resolution X-ray detectors

    International Nuclear Information System (INIS)

    Hess, U.; Buehler, M.; Hentig, R. von; Hertrich, T.; Phelan, K.; Wernicke, D.; Hoehne, J.

    2001-01-01

    CSP Cryogenic Spectrometers GmbH is developing cryogenic energy dispersive X-ray spectrometers based on superconducting detector technology. Superconducting sensors exhibit at least a 10-fold improvement in energy resolution due to their low energy gap compared to conventional Si(Li) or Ge detectors. These capabilities are extremely valuable for the analysis of light elements and in general for the analysis of the low energy range of the X-ray spectrum. The spectrometer is based on a mechanical cooler needing no liquid coolants and an adiabatic demagnetization refrigerator (ADR) stage which supplies the operating temperature of below 100 mK for the superconducting sensor. Applications include surface analysis in semiconductor industry as well material analysis for material composition e.g. in ceramics or automobile industry

  6. Flight performance of an advanced CZT imaging detector in a balloon-borne wide-field hard X-ray telescope-ProtoEXIST1

    Energy Technology Data Exchange (ETDEWEB)

    Hong, J., E-mail: jaesub@head.cfa.harvard.edu [Harvard-Smithsonian Center for Astrophysics, Cambridge, MA 02138 (United States); Allen, B.; Grindlay, J. [Harvard-Smithsonian Center for Astrophysics, Cambridge, MA 02138 (United States); Barthelemy, S.; Baker, R. [NASA Goddard Space Flight Center, Greenbelt, MD 20771 (United States); Garson, A.; Krawczynski, H. [Washington University in St. Louis and the McDonnell Center for the Space Sciences, St. Louis, MO 63130 (United States); Apple, J.; Cleveland, W.H. [NASA Marshall Space Flight Center and Universities Space Research Association, Huntsville, AL 35812 (United States)

    2011-10-21

    We successfully carried out the first high-altitude balloon flight of a wide-field hard X-ray coded-aperture telescope ProtoEXIST1, which was launched from the Columbia Scientific Balloon Facility at Ft. Sumner, New Mexico on October 9, 2009. ProtoEXIST1 is the first implementation of an advanced CdZnTe (CZT) imaging detector in our ongoing program to establish the technology required for next generation wide-field hard X-ray telescopes such as the High Energy Telescope (HET) in the Energetic X-ray Imaging Survey Telescope (EXIST). The CZT detector plane in ProtoEXIST1 consists of an 8x8 array of closely tiled 2 cmx2 cmx0.5 cm thick pixellated CZT crystals, each with 8x8 pixels, mounted on a set of readout electronics boards and covering a 256 cm{sup 2} active area with 2.5 mm pixels. A tungsten mask, mounted at 90 cm above the detector provides shadowgrams of X-ray sources in the 30-600 keV band for imaging, allowing a fully coded field of view of 9{sup o}x9{sup o} (and 19{sup o}x19{sup o} for 50% coding fraction) with an angular resolution of 20'. In order to reduce the background radiation, the detector is surrounded by semi-graded (Pb/Sn/Cu) passive shields on the four sides all the way to the mask. On the back side, a 26 cmx26 cmx2 cm CsI(Na) active shield provides signals to tag charged particle induced events as well as {>=}100keV background photons from below. The flight duration was only about 7.5 h due to strong winds (60 knots) at float altitude (38-39 km). Throughout the flight, the CZT detector performed excellently. The telescope observed Cyg X-1, a bright black hole binary system, for {approx}1h at the end of the flight. Despite a few problems with the pointing and aspect systems that caused the telescope to track about 6.4{sup o} off the target, the analysis of the Cyg X-1 data revealed an X-ray source at 7.2{sigma} in the 30-100 keV energy band at the expected location from the optical images taken by the onboard daytime star camera. The

  7. X-ray detector for a panoramic X-ray unit

    Energy Technology Data Exchange (ETDEWEB)

    Cowell, D; Ensslin, F H

    1976-01-15

    The discovery deals with an X-ray detector suitable for the controlling of panoramic X-ray systems. It consists of a fluorescent image screen and a semiconductor photo cell. The output signal of the detector is proportional to the intensity of the X-radiation and the response time is large enough to follow the change of amplitude of the contours of the modulated X radiation. The detector with band-pass filter regulates, via a control system, the moving rate of the X-ray source and of the film opposite it in dependence of the intensity, so that a uniform exposure is ensured.

  8. High-Resolution Detector For X-Ray Diffraction

    Science.gov (United States)

    Carter, Daniel C.; Withrow, William K.; Pusey, Marc L.; Yost, Vaughn H.

    1988-01-01

    Proposed x-ray-sensitive imaging detector offers superior spatial resolution, counting-rate capacity, and dynamic range. Instrument based on laser-stimulated luminescence and reusable x-ray-sensitive film. Detector scans x-ray film line by line. Extracts latent image in film and simultaneously erases film for reuse. Used primarily for protein crystallography. Principle adapted to imaging detectors for electron microscopy and fluorescence spectroscopy and general use in astronomy, engineering, and medicine.

  9. X-ray detectors based on image sensors

    International Nuclear Information System (INIS)

    Costa, A.P.R.

    1983-01-01

    X-ray detectors based on image sensors are described and a comparison is made between the advantages and the disadvantages of such a kind of detectors with the position sensitive detectors. (L.C.) [pt

  10. The one- and two-coordinate x-ray detectors

    International Nuclear Information System (INIS)

    Aulchenko, V.M.; Baru, S.E.; Khabakhpashev, A.G.; Savinov, G.A.

    1992-01-01

    The Institute of Nuclear Physics has designed and fabricated one- and two-coordinate x-ray detectors since 1975. For photon detection multiwire proportional chambers that operate in direct pulse count mode are employed. The characteristics of the detectors allow successful use of them for a wide range of diffractive x-ray structure studies, including studies of dynamics of structure variation (x-ray diffractive movies) and measurements at synchrotron radiation channels

  11. Small area silicon diffused junction X-ray detectors

    Science.gov (United States)

    Walton, J. T.; Pehl, R. H.; Larsh, A. E.

    1982-01-01

    The low-temperature performance of silicon diffused junction detectors in the measurement of low energy X-rays is reported. The detectors have an area of 0.04 sq cm and a thickness of 100 microns. The spectral resolutions of these detectors were found to be in close agreement with expected values, indicating that the defects introduced by the high-temperature processing required in the device fabrication were not deleteriously affecting the detection of low-energy X-rays. Device performance over a temperature range of 77 K to 150 K is given. These detectors were designed to detect low-energy X-rays in the presence of minimum ionizing electrons. The successful application of silicon-diffused junction technology to X-ray detector fabrication may facilitate the development of other novel silicon X-ray detector designs.

  12. Small area silicon diffused junction x-ray detectors

    International Nuclear Information System (INIS)

    Walton, J.T.; Pehl, R.H.; Larsh, A.E.

    1981-10-01

    The low temperature performance of silicon diffused junction detectors in the measurement of low energy x-rays is reported. The detectors have an area of 0.04 cm 2 and a thickness of 100 μm. The spectral resolutions of these detectors were found to be in close agreement with expected values indicating that the defects introduced by the high temperature processing required in the device fabrication were not deleteriously affecting the detection of low energy x-rays. Device performance over a temperature range of 77 to 150 0 K is given. These detectors were designed to detect low energy x-rays in the presence of minimum ionizing electrons. The successful application of silicon diffused junction technology to x-ray detector fabrication may facilitate the development of other novel silicon x-ray detector designs

  13. CCD-based X-ray detectors for X-ray diffraction studies

    International Nuclear Information System (INIS)

    Ito, K.; Amemiya, Y.

    1999-01-01

    CCD-based X-ray detectors are getting to be used for X-ray diffraction studies especially in the studies where real time (automated) measurements and time-resolved measurements are required. Principles and designs of two typical types of CCD-based detectors are described; one is ths system in which x-ray image intensifiers are coupled to maximize the detective quantum efficiency for time-resolved measurements, and the other is the system in which tapered optical fibers are coupled for the reduction of the image into the CCD, which is optimized for automated measurements for protein crystallography. These CCD-based X-ray detectors have an image distortion and non-uniformity of response to be corrected by software. Correction schemes which we have developed are also described. (author)

  14. Small Pixel Hybrid CMOS X-ray Detectors

    Science.gov (United States)

    Hull, Samuel; Bray, Evan; Burrows, David N.; Chattopadhyay, Tanmoy; Falcone, Abraham; Kern, Matthew; McQuaide, Maria; Wages, Mitchell

    2018-01-01

    Concepts for future space-based X-ray observatories call for a large effective area and high angular resolution instrument to enable precision X-ray astronomy at high redshift and low luminosity. Hybrid CMOS detectors are well suited for such high throughput instruments, and the Penn State X-ray detector lab, in collaboration with Teledyne Imaging Sensors, has recently developed new small pixel hybrid CMOS X-ray detectors. These prototype 128x128 pixel devices have 12.5 micron pixel pitch, 200 micron fully depleted depth, and include crosstalk eliminating CTIA amplifiers and in-pixel correlated double sampling (CDS) capability. We report on characteristics of these new detectors, including the best read noise ever measured for an X-ray hybrid CMOS detector, 5.67 e- (RMS).

  15. Charge collection and absorption-limited x-ray sensitivity of pixellated x-ray detectors

    International Nuclear Information System (INIS)

    Kabir, M. Zahangir; Kasap, S.O.

    2004-01-01

    The charge collection and absorption-limited x-ray sensitivity of a direct conversion pixellated x-ray detector operating in the presence of deep trapping of charge carriers is calculated using the Shockley-Ramo theorem and the weighting potential of the individual pixel. The sensitivity of a pixellated x-ray detector is analyzed in terms of normalized parameters; (a) the normalized x-ray absorption depth (absorption depth/photoconductor thickness), (b) normalized pixel width (pixel size/thickness), and (c) normalized carrier schubwegs (schubweg/thickness). The charge collection and absorption-limited sensitivity of pixellated x-ray detectors mainly depends on the transport properties (mobility and lifetime) of the charges that move towards the pixel electrodes and the extent of dependence increases with decreasing normalized pixel width. The x-ray sensitivity of smaller pixels may be higher or lower than that of larger pixels depending on the rate of electron and hole trapping and the bias polarity. The sensitivity of pixellated detectors can be improved by ensuring that the carrier with the higher mobility-lifetime product is drifted towards the pixel electrodes

  16. X-ray imaging with the PILATUS 100k detector

    DEFF Research Database (Denmark)

    Bech, Martin; Bunk, O.; David, C.

    2008-01-01

    We report on the application of the PILATUS 100K pixel detector for medical imaging. Experimental results are presented in the form of X-ray radiographs using standard X-ray absorption contrast and a recently developed phase contrast imaging method. The results obtained with the PILATUS detector...... are compared to results obtained with a conventional X-ray imaging system consisting of an X-ray scintillation screen, lens optics, and a charge coupled device. Finally, the results for both systems are discussed more quantitatively based on an image power spectrum analysis. Udgivelsesdato: April...

  17. The color of X-rays: Spectral X-ray computed tomography using energy sensitive pixel detectors

    NARCIS (Netherlands)

    Schioppa, E.J.

    2014-01-01

    Energy sensitive X-ray imaging detectors are produced by connecting a semiconductor sensor to a spectroscopic pixel readout chip. In this thesis, the applicability of such detectors to X-ray Computed Tomography (CT) is studied. A prototype Medipix based silicon detector is calibrated using X-ray

  18. Soft x-ray detection with diamond photoconductive detectors

    International Nuclear Information System (INIS)

    Kania, D.R.; Pan, L.; Kornblum, H.; Bell, P.; Landen, O.N.; Pianetta, P.

    1990-01-01

    Photoconductive detectors fabricated from natural lla diamonds have been used to measure the x-ray power emitted from laser produced plasmas. The detector was operated without any absorbing filters to distort the x-ray power measurement. The 5.5 eV bandgap of the detector material practically eliminates its sensitivity to scattered laser radiation thus permitting filterless operation. The detector response time or carrier life time was 90 ps. Excellent agreement was achieved between a diamond PCD and a multichannel photoemissive diode array in the measurement of radiated x-ray power and energy. 4 figs

  19. Medical imaging: Material change for X-ray detectors

    Science.gov (United States)

    Rowlands, John A.

    2017-10-01

    The X-ray sensitivity of radiology instruments is limited by the materials used in their detectors. A material from the perovskite family of semiconductors could allow lower doses of X-rays to be used for medical imaging. See Letter p.87

  20. Development of X-ray detector based on phototransistor

    International Nuclear Information System (INIS)

    Ramacos Fardela; Kusminarto

    2014-01-01

    X-ray interaction with matter can produce phenomenon of fluorescence that emits visible light. This phenomenon has been exploited to design an X-ray detector based on photo transistor by attaching a screen ZnS(Ag) on the surface of the photo transistor which is arranged in a Darlington circuit. Response of detector was done by collimating of X-rays beam from the X-ray generator tube Philips 2000 watts, 60 kV type PW 2215/20 NR 780 026 and measure the detector output voltage (V out ). Varying the current by 5, 10, 15, 20, 25, 30, 35 and 40 mA in the X-ray panel. The experimental results showed that the Darlington circuit can be applied to design the detector of X-ray based on phototransistor. The results show that there is a linear relationship between the change in the intensity of X-ray detectors with voltage output phototransistor when it was closed with fluorescence materials ZnS(Ag), the linearity coefficient was R 2 = 0.99. Sensitivity of detector was obtained to be 3.7 x 10 -2 mV per cpm. (author)

  1. Gamma detector for use with luggage X-ray systems

    International Nuclear Information System (INIS)

    McHugh, H.; Quam, W.

    1998-01-01

    A new gamma radiation sensor has been designed for installation on several types of luggage x-ray machines and mobile x-ray vans operated by the U.S. Customs Service and the U.S. Department of State. The use of gamma detectors on x-ray machines imposed difficulties not usually encountered in the design of gamma detectors because the spectrum of scattered x-rays, which varied from machine to machine, extended to energies significantly higher than those of the low-energy isotopic emissions. In the original design, the lower level discriminator was raised above the x-ray end point energy resulting in the loss of the americium line associated with plutonium. This reduced the overall sensitivity to unshielded plutonium by a factor of approximately 100. An improved method was subsequently developed wherein collimation was utilized in conjunction with a variable counting threshold to permit accommodation of differing conditions of x-ray scattering. This design has been shown to eliminate most of the problems due to x-ray scattering while still capturing the americium emissions. The overall sensitivity has remained quite high, though varying slightly from one model of x-ray machine to another, depending upon the x-ray scattering characteristics of each model. (author)

  2. Radiation detectors of PIN type for X-rays

    International Nuclear Information System (INIS)

    Ramirez-Jimenez, F.J.

    2003-01-01

    In this laboratory session, tree experiments are proposed: the measurement of X-ray energy spectra from radioactive sources with a high resolution cooled Si-Li detector, with a room temperature PIN diode and the measurement of the response of a PIN diode to the intensity of X-rays of radio-diagnostic units. The spectra obtained with the Si-Li detector help to understand the energy distribution of X-rays and are used as a reference to compare the results obtained with the PIN diode. Measurements in medical X-ray machines are proposed. Low cost, simple electronic instruments and systems are used as tools to make measurements in X-ray units used in radio-diagnostic

  3. Ultrafast secondary emission x-ray imaging detectors

    International Nuclear Information System (INIS)

    Akkerman, A.; Gibrekhterman, A.; Majewski, S.

    1991-07-01

    Fast high accuracy, x-ray imaging at high photon flux can be achieved when coupling thin solid convertors to gaseous electron multipliers, operating at low gas pressures. Secondary electron emitted from the convertor foil are multiplied in several successive amplification elements. The obvious advantage of solid x-ray detectors, as compared to gaseous conversion, are the production of parallax-free images and the fast (subnanoseconds) response. These x-ray detectors have many potential applications in basic and applied research. Of particular interest is the possibility of an efficient and ultrafast high resolution imaging of transition radiation,with a reduced dE/dx background. We present experimental results on the operation of the secondary emission x-ray (SEX) detectors, their detection efficiency, localization and time resolution. The experimental work is accompanied by mathematical modelling and computer simulation of transition radiation detectors based on CsI transition radiation convertors. (author)

  4. Gas detectors for x-ray lasers

    International Nuclear Information System (INIS)

    Tiedtke, K.; Feldhaus, J.; Hahn, U.; Jastrow, U.; Nunez, T.; Tschentscher, T.; Bobashev, S. V.; Sorokin, A. A.; Hastings, J. B.; Moeller, S.; Cibik, L.; Gottwald, A.; Hoehl, A.; Kroth, U.; Krumrey, M.; Schoeppe, H.; Ulm, G.; Richter, M.

    2008-01-01

    We have developed different types of photodetectors that are based on the photoionization of a gas at a low target density. The almost transparent devices were optimized and tested for online photon diagnostics at current and future x-ray free-electron laser facilities on a shot-to-shot basis with a temporal resolution of better than 100 ns. Characterization and calibration measurements were performed in the laboratory of the Physikalisch-Technische Bundesanstalt at the electron storage ring BESSY II in Berlin. As a result, measurement uncertainties of better than 10% for the photon-pulse energy and below 20 μm for the photon-beam position were achieved at the Free-electron LASer in Hamburg (FLASH). An upgrade for the detection of hard x-rays was tested at the Sub-Picosecond Photon Source in Stanford

  5. Gas detectors for x-ray lasers

    Science.gov (United States)

    Tiedtke, K.; Feldhaus, J.; Hahn, U.; Jastrow, U.; Nunez, T.; Tschentscher, T.; Bobashev, S. V.; Sorokin, A. A.; Hastings, J. B.; Möller, S.; Cibik, L.; Gottwald, A.; Hoehl, A.; Kroth, U.; Krumrey, M.; Schöppe, H.; Ulm, G.; Richter, M.

    2008-05-01

    We have developed different types of photodetectors that are based on the photoionization of a gas at a low target density. The almost transparent devices were optimized and tested for online photon diagnostics at current and future x-ray free-electron laser facilities on a shot-to-shot basis with a temporal resolution of better than 100 ns. Characterization and calibration measurements were performed in the laboratory of the Physikalisch-Technische Bundesanstalt at the electron storage ring BESSY II in Berlin. As a result, measurement uncertainties of better than 10% for the photon-pulse energy and below 20 μm for the photon-beam position were achieved at the Free-electron LASer in Hamburg (FLASH). An upgrade for the detection of hard x-rays was tested at the Sub-Picosecond Photon Source in Stanford.

  6. Hight resolution Si(Li) X ray detector

    International Nuclear Information System (INIS)

    Yuan Xianglin; Huang Naizhang; Lin Maocai; Li Zhiyong

    1985-01-01

    This paper describes the fabrication technology of GL1221 type Si(Li) X ray detector core and the pulse light feedback colded preamplifier fitted on the detector. The energy resolution of the detector system is 165 eV (At 5.89 KeV Mn-K α X ray); the counting rate is 1020 cps, and the electronics noise is 104 eV. The performace of the detector keeps up with the business level of a foreign product of the same kind

  7. Position-sensitive X-ray detectors

    International Nuclear Information System (INIS)

    Hendrix, J.

    1982-01-01

    In this review of the application of different types of position sensitive detectors to synchrotron radiation, discussion of the proportional counters based on the gas amplification principle forms a major part. Other topics reviewed are detector requirements, multiwire proportional chamber system, drift chamber type detectors, TV detectors, and recent developments, such as that based on a micro-channel plate as the amplifying element, and charge-coupled devices. (U.K.)

  8. Gated x-ray detector for the National Ignition Facility

    International Nuclear Information System (INIS)

    Oertel, John A.; Aragonez, Robert; Archuleta, Tom; Barnes, Cris; Casper, Larry; Fatherley, Valerie; Heinrichs, Todd; King, Robert; Landers, Doug; Lopez, Frank; Sanchez, Phillip; Sandoval, George; Schrank, Lou; Walsh, Peter; Bell, Perry; Brown, Matt; Costa, Robert; Holder, Joe; Montelongo, Sam; Pederson, Neal

    2006-01-01

    Two new gated x-ray imaging cameras have recently been designed, constructed, and delivered to the National Ignition Facility in Livermore, CA. These gated x-Ray detectors are each designed to fit within an aluminum airbox with a large capacity cooling plane and are fitted with an array of environmental housekeeping sensors. These instruments are significantly different from earlier generations of gated x-ray images due, in part, to an innovative impedance matching scheme, advanced phosphor screens, pulsed phosphor circuits, precision assembly fixturing, unique system monitoring, and complete remote computer control. Preliminary characterization has shown repeatable uniformity between imaging strips, improved spatial resolution, and no detectable impedance reflections

  9. Fabrication of Sparse Readout Detectors for X-ray Astronomy

    Science.gov (United States)

    Burrows, David

    We propose to continue our detector development program in X-ray astronomy. Under our current APRA grant we have fabricated a new read out integrated circuit that is one half of a hybrid CMOS detector. Here we propose to build and test these innovative detectors, which could potentially be flown on future X-ray missions with focused optics and/or large effective area. This proposal supports NASA's goals of technical advancement of technologies suitable for future missions and training of graduate students.

  10. Applications for X-ray detectors in astrophysics

    International Nuclear Information System (INIS)

    Remillard, R.A.

    2003-01-01

    Full text: Position-sensitive X-Ray detectors continue to playa central role in high-energy astrophysics. The current science goals are reviewed with emphasis on requirements in terms of camera performance. Wide-field imaging techniques, including coded mask cameras, are an essential part of space programs because of the transient nature of high-priority targets, e.g. eruptions from black-hole binaries and cosmic explosions such as gamma ray bursts. Pointing X-ray telescopes are being planned with a wide range of photon energies and with collection designs that include both mirrors and coded masks. Requirements for high spectral resolution and high time resolution are driven by diverse types of X-ray sources such as msec pulsars, quasars with emission-line profiles shaped by general relativity, and X-ray binaries that exhibit quasi-periodic oscillations in the range of 40-1300 Hz. Many laboratories and universities are involved in space-qualification of new detector technologies, e.g. CZT cameras, X-ray calorimeters, new types of CCDs, and GEM detectors. Even X-ray interferometry is on the horizon of NASA's science roadmap. The difficulties in advancing new technologies for space science applications require careful coordinations between industry and science groups in order to solve science problems while minimizing risk

  11. An x-ray detector for time-resolved studies

    International Nuclear Information System (INIS)

    Rodricks, B.; Brizard, C.; Clarke, R.; Lowe, W.

    1992-01-01

    The development of ultrahigh-brightness x-ray sources makes time-resolved x-ray studies more and more feasible. Improvements in x-ray optics components are also critical for obtaining the appropriate beam for a particular type of experiment. Moreover, fast parallel detectors will be essential in order to exploit the combination of high intensity x-ray sources and novel optics for time-resolved experiments. A CCD detector with a time resolution of microseconds has been developed at the Advanced Photon Source (APS). This detector is fully programmable using CAMAC electronics and a Micro Vax computer. The techniques of time-resolved x-ray studies, which include scattering, microradiography, microtomography, stroboscopy, etc., can be applied to a range of phenomena (including rapid thermal annealing, surface ordering, crystallization, and the kinetics of phase transition) in order to understand these time-dependent microscopic processes. Some of these applications will be illustrated by recent results performed at synchrotrons. New powerful x-ray sources now under construction offer the opportunity to apply innovative approaches in time-resolved work

  12. The digital flat-panel X-Ray detectors

    International Nuclear Information System (INIS)

    Risticj, S. Goran

    2013-01-01

    In a digital imaging system, the incident x-ray image must be sampled both in the spatial and intensity dimensions. In the spatial dimensions, samples are obtained as averages of the intensity over picture elements or pixels. In the intensity dimension, the signal is digitalized into one of a finite number of levels or bits. Two main types of digital flat-panel detectors are based on the direct conversion, which contains the photoconductor, and on indirect conversion, which contains phosphor. The basics of these detectors are given. Coupling traditional x-ray detection material such as photoconductors and phosphors with a large-area active-matrix readout structure forms the basis of flat panel x-ray images. Active matrix technology provides a new, highly efficient, real time method for electronically storing and measuring the product of the x-ray interaction stage whether the product is visible wavelength photons or electrical charges. The direct and indirect detectors, made as the active-matrix flat-panel detectors containing sensing/storage elements, switching elements (diodes or thin film transistors (TFTS)) and image processing module, are described. Strengths and limitations of stimulable phosphors are discussed. The main advantages and disadvantages of mentioned x-ray detectors are also analyzed. (Author)

  13. A gas pixel detector for X-ray imaging

    International Nuclear Information System (INIS)

    Bateman, J.E.; Connolly, J.F.

    1991-11-01

    A simple, robust form of gas pixel detector is discussed which is based on the use of electronic connector pins as the gain elements. With a rate capability of >10 5 counts/s per pin, an X-ray imaging detector system capable of counting at global rates of the order of 10 10 counts/s is foreseen. (author)

  14. Modeling and design of X-rays bidimensional detectors

    International Nuclear Information System (INIS)

    Quisbert, Elmer Paz Alcon

    2000-03-01

    In this work has been developed the scintillating fiber optic and semiconductor devices based 2-D detector design, modeling and performance evaluation using Monte Carlo methods, for high X-ray energy range (10-140 kV) radiography and tomography applications. These processes allowed us, also, the imaging system parameters and components optimization and appropriate detector design. The model estimated the detectors performance parameters (DQE, MTF and SNR), and radiation risk (in terms of mean absorbed dose in the patient) and to show up how the sequence of physical processes in X-ray detection influence the performance of this imaging PFOC detectors. In this way, the modeling of the detector includes the statistics of the spatial distribution of absorbed X-rays and of X-ray to light conversion, its transmission, and the light quanta conversion into electrons. Also contributions to noise from the detection system chain is included, mainly the CCD detector ambient noise. Performance prediction, based on calculation taken from simulations, illustrates how such detectors meet the exacting requirements of some medical and industrial applications. Also, it is envisaged that our modeling procedure of the imaging system will be suitable not only for investigating how the system components should be best designed but for CT and RD system performance prediction. The powerful techniques would enable us to give advice for future development, in this field, in search of more dose-efficient imaging systems. (author)

  15. Development of an X-ray detector using photodiodes

    International Nuclear Information System (INIS)

    Gonzalez G, J.; Azorin V, J. C.; Sosa A, M. A.; Ceron, P.

    2016-10-01

    Currently the radiation detectors for medical applications are very high value in the market and are difficult to access as training material. In the Sciences and Engineering Division of the Guanajuato University (Mexico) investigations are carried out related to ionizing radiations, especially with X-rays. To overcome the lack of materials has had to resort to borrowing equipment from other institutions, so its use and availability are intermittent. For these reasons is proposed to design and implement an X-ray detector for the use of the work group and the University. This work aims to build an X-ray semiconductor detector using inexpensive and affordable materials, is also proposed the use of a photodiode sensor and an Arduino analog-digital card and a LCD display showing the data. (Author)

  16. A multiple CCD X-ray detector and its first operation with synchrotron radiation X-ray beam

    CERN Document Server

    Suzuki, M; Kumasaka, T; Sato, K; Toyokawa, H; Aries, I F; Jerram, P A; Ueki, T

    1999-01-01

    A 4x4 array structure of 16 identical CCD X-ray detector modules, called the multiple CCD X-ray detector system (MCCDX), was submitted to its first synchrotron radiation experiment at the protein crystallography station of the RIKEN beamline (BL45XU) at the SPring-8 facility. An X-ray diffraction pattern of cholesterol powder was specifically taken in order to investigate the overall system performance.

  17. Study of semiconductor detectors applied to diagnostic X-ray

    International Nuclear Information System (INIS)

    Salgado, Cesar Marques

    2003-08-01

    This work aims an evaluation of procedures for photons spectrum determination, produced by a X ray tube, normally used for medical diagnoses which operation voltage ranges from 20 to 150 kVp, to allow more precise characterization of the photon beam. The use of spectrum analysis will contribute to reduce the uncertainty in the ionization camera calibrations. For this purpose, two kind of detectors were selected, a Cadmium Zinc Telluride (CZT) and a planar HPGe detector. The X ray interaction with the detector's crystal produces, by electronic processes, a pulse high distribution as an output, which is no the true photon spectrum, due to the presence of K shell escape peaks, Compton scattering and to the fact that the detectors efficiency diminish rapidly with the increase of the photon energy. A detailed analysis of the contributing factors to distortions in the spectrum is necessary and was performed by Monte Carlo calculation with the MCNP 4B computer code. In order to determine the actual photon spectrum for a X ray tube a spectra stripping procedure is described for the HPGe detector. The detector's response curves, determined by the Monte Carlo calculation, were compared to the experimental ones, for isotropic point sources. For the methodology validation, stripped spectra were compared to the theoretical ones, for the same X ray tube's settings, for a qualitative evaluation. The air kerma rate calculated with the photon spectra were compared to the direct measurement using an ionization chamber, for a quantitative evaluation. (author)

  18. Flat-panel detectors in x-ray diagnosis

    International Nuclear Information System (INIS)

    Spahn, M.; Heer, V.; Freytag, R.

    2003-01-01

    For all application segments X-ray systems with flat-panel detectors increasingly enter the market. In digital radiography, mammography and cardiologic angiography flat-panel detectors are already well established while they are made ready for market introduction in general angiography and fluoroscopy. Two flat-panel detector technologies are available. One technology is based on an indirect conversion process of X-rays while the other one uses a direct conversion method.For radiography and dynamic applications the indirect method provides substantial advantages, while the direct method has some benefits for mammography. In radiography and mammography flat-panel detectors lead to clear improvements with respect to workflow, image quality and dose reduction potentials. These improvements are fostered by the immediate availability of the image, the large dynamic range and the high sensitivity to X-rays. New applications and the use of complex image processing algorithms have the potential to enlarge the present diagnostic range of applications.Up to now, image intensifiers are still the well-established technology for angiography and fluoroscopy. Nevertheless flat-panel detectors begin to enter this field, especially in cardiologic angiography.Characteristics of flat-panel detectors such as the availability of distortion-free images, the excellent contrast resolution, the large dynamic range, the high sensitivity to X-rays and the usability in magnetic fields provide the basis for improved and new diagnostic and interventional methods. (orig.) [de

  19. X-ray imaging bilinear staggered GaAs detectors

    Energy Technology Data Exchange (ETDEWEB)

    Achmadullin, R.A.; Dvoryankin, V.F. E-mail: vfd217@ire216.msk.su; Dvoryankina, G.G.; Dikaev, Y.M.Yu.M.; Krikunov, A.I.; Kudryashov, A.A.; Panova, T.M.; Petrov, A.G.; Telegin, A.A

    2004-09-21

    The multichannel bilinear X-ray detector based on epitaxial GaAs structures is developed to obtain a digital X-ray image. Each detector operates in photovoltaic mode without reverse bias that enables almost complete elimination of detector noise arising due to leakage currents. The sensitivity range of the epitaxial GaAs photovoltaic X-ray detector covers the effective energies from 8 to 120 keV. A maximum response of the detector operating in the short-circuit mode was observed at an energy of 35 keV and amounted to 30 {mu}A min/(Gy cm{sup 2}). The multichannel detector was made of 1024 pixels with pitch of 0.8 mm. The spatial resolution of double staggered sensor row is twice as high as the resolution of that of single sensor row with the same pitch. Measured spatial resolution is 1.2 line-pairs/mm, contrast sensitivity not worse 1% and dynamic range defined as the ratio of maximum detectable X-ray signal to electronic noise level more than 2000 are received.

  20. X-ray imaging bilinear staggered GaAs detectors

    International Nuclear Information System (INIS)

    Achmadullin, R.A.; Dvoryankin, V.F.; Dvoryankina, G.G.; Dikaev, Y.M.Yu.M.; Krikunov, A.I.; Kudryashov, A.A.; Panova, T.M.; Petrov, A.G.; Telegin, A.A.

    2004-01-01

    The multichannel bilinear X-ray detector based on epitaxial GaAs structures is developed to obtain a digital X-ray image. Each detector operates in photovoltaic mode without reverse bias that enables almost complete elimination of detector noise arising due to leakage currents. The sensitivity range of the epitaxial GaAs photovoltaic X-ray detector covers the effective energies from 8 to 120 keV. A maximum response of the detector operating in the short-circuit mode was observed at an energy of 35 keV and amounted to 30 μA min/(Gy cm 2 ). The multichannel detector was made of 1024 pixels with pitch of 0.8 mm. The spatial resolution of double staggered sensor row is twice as high as the resolution of that of single sensor row with the same pitch. Measured spatial resolution is 1.2 line-pairs/mm, contrast sensitivity not worse 1% and dynamic range defined as the ratio of maximum detectable X-ray signal to electronic noise level more than 2000 are received

  1. Tutorial on X-ray photon counting detector characterization.

    Science.gov (United States)

    Ren, Liqiang; Zheng, Bin; Liu, Hong

    2018-01-01

    Recent advances in photon counting detection technology have led to significant research interest in X-ray imaging. As a tutorial level review, this paper covers a wide range of aspects related to X-ray photon counting detector characterization. The tutorial begins with a detailed description of the working principle and operating modes of a pixelated X-ray photon counting detector with basic architecture and detection mechanism. Currently available methods and techniques for charactering major aspects including energy response, noise floor, energy resolution, count rate performance (detector efficiency), and charge sharing effect of photon counting detectors are comprehensively reviewed. Other characterization aspects such as point spread function (PSF), line spread function (LSF), contrast transfer function (CTF), modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), bias voltage, radiation damage, and polarization effect are also remarked. A cadmium telluride (CdTe) pixelated photon counting detector is employed for part of the characterization demonstration and the results are presented. This review can serve as a tutorial for X-ray imaging researchers and investigators to understand, operate, characterize, and optimize photon counting detectors for a variety of applications.

  2. Development of multiwire gas detectors for X-rays

    International Nuclear Information System (INIS)

    Sales, Eraldo de

    2015-01-01

    This work presents the prototype of a 2D position sensitive gas detector for application in X-ray scattering and diffraction experiments. Starting from a detector initially developed for other applications and will show the required changes on the original concept of this device. The strategy used to determine the necessary adaptations were based on searching in the literature for the overall characteristics of a multi-wire X-ray detector (choice of gas, pressure, window, etc.), the use of simulations, implementation of the changes and finally operational tests. Computational tools were used to calculate the mechanical strength and attenuation of the X-ray photons that helped to determine the most appropriate material for the construction of the entrance window. Detector simulations were built with Garfield software and were used to study the overall properties of the detector, and to determine the optimum parameters for the equipment operation. Typical parameters are the distance between the wires, wire diameter, high voltage to be used, among several other parameters. The results obtained showed that the multi-wire detector concept with the implemented adaptations allowed the detection of X-rays. However, depending on the application, it may be necessary improve the resolution of the equipment, in order to have a better description of the collected data. Several ideas are suggested for this improvement. It is also presented interesting results obtained with a microscopic pattern detector called triple GEM. This device belongs to the Gas Detectors Development group (GDD group) at CERN and was used in my training at this institution. The results showed the potential of the equipment for detection of X-rays. The results and simulations presented in this work, confirmed that the changes in the concept of the original detector permitted it use on X-ray detection applications. Also, it was possible to obtain several indications for further optimization, which may

  3. Gas Pixel Detectors for low energy X-ray polarimetry

    International Nuclear Information System (INIS)

    Spandre, Gloria

    2007-01-01

    Gas Pixel Detectors are position-sensitive proportional counters in which a complete integration between the gas amplification structure and the read-out electronics has been reached. Various generation of Application-Specific Integrated Circuit (ASIC) have been designed in deep submicron CMOS technology to realize a monolithic device which is at the same time the charge collecting electrode and the analog amplifying and charge measuring front-end electronics. The experimental response of a detector with 22060 pixels at 80 μm pitch to polarized and un-polarized X-ray radiation is shown and the application of this device for Astronomical X-ray Polarimetry discussed

  4. Recent X-ray hybrid CMOS detector developments and measurements

    Science.gov (United States)

    Hull, Samuel V.; Falcone, Abraham D.; Burrows, David N.; Wages, Mitchell; Chattopadhyay, Tanmoy; McQuaide, Maria; Bray, Evan; Kern, Matthew

    2017-08-01

    The Penn State X-ray detector lab, in collaboration with Teledyne Imaging Sensors (TIS), have progressed their efforts to improve soft X-ray Hybrid CMOS detector (HCD) technology on multiple fronts. Having newly acquired a Teledyne cryogenic SIDECARTM ASIC for use with HxRG devices, measurements were performed with an H2RG HCD and the cooled SIDECARTM. We report new energy resolution and read noise measurements, which show a significant improvement over room temperature SIDECARTM operation. Further, in order to meet the demands of future high-throughput and high spatial resolution X-ray observatories, detectors with fast readout and small pixel sizes are being developed. We report on characteristics of new X-ray HCDs with 12.5 micron pitch that include in-pixel CDS circuitry and crosstalk-eliminating CTIA amplifiers. In addition, PSU and TIS are developing a new large-scale array Speedster-EXD device. The original 64 × 64 pixel Speedster-EXD prototype used comparators in each pixel to enable event driven readout with order of magnitude higher effective readout rates, which will now be implemented in a 550 × 550 pixel device. Finally, the detector lab is involved in a sounding rocket mission that is slated to fly in 2018 with an off-plane reflection grating array and an H2RG X-ray HCD. We report on the planned detector configuration for this mission, which will increase the NASA technology readiness level of X-ray HCDs to TRL 9.

  5. Thermally stimulated investigations on diamond X-Ray detectors

    International Nuclear Information System (INIS)

    Tromson, D.; Bergonzo, P.; Brambilla, A.; Mer, C.; Foulon, F.; Amosov, V.N.

    1999-01-01

    Intrinsic diamond material is increasingly used for the fabrication of radiation detectors. However, the presence of inherent defects has a strong impact on the detector characteristics such as the time dependent stability of the detection signal. In order to draw better insights into this effect, comparative investigations of the X-ray responses with thermally stimulated current (TSC) measurements were carried out on natural diamond detectors. TSC revealed the presence of four peaks or shoulders on natural samples in the 200 to 500 K domain. Three energy levels were identified at about 0.7, 0.71 and 0.95 eV. Time dependent X-ray detector sensitivity was investigated for various initial conditions. The results give evidence of the improvement of the detection properties after having filled traps in the material by X-ray irradiation. The comparison between the X-ray response and the TSC spectra indicate that trapping levels emptied at room temperature appear to significantly affect the performance of radiation detectors. (authors)

  6. X-Ray and Gamma-Ray Radiation Detector

    DEFF Research Database (Denmark)

    2015-01-01

    Disclosed is a semiconductor radiation detector for detecting X-ray and / or gamma-ray radiation. The detector comprises a converter element for converting incident X-ray and gamma-ray photons into electron-hole pairs, at least one cathode, a plurality of detector electrodes arranged with a pitch...... (P) along a first axis, a plurality of drift electrodes, a readout circuitry being configured to read out signals from the plurality of detector electrodes and a processing unit connected to the readout circuitry and being configured to detect an event in the converter element. The readout circuitry...... is further configured to read out signals from the plurality of drift electrodes, and the processing unit is further configured to estimate a location of the event along the first axis by processing signals obtained from both the detector electrodes and the drift electrodes, the location of the event along...

  7. Diamond detectors for synchrotron radiation X-ray applications

    Energy Technology Data Exchange (ETDEWEB)

    De Sio, A. [Laboratori Nazionali di Frascati, INFN, 00044 Frascati, Roma (Italy); Department of Astronomy and Space Science, Universita di Firenze, L.go E. Fermi 2, 50125 Firenze (Italy)], E-mail: desio@arcetri.astro.it; Pace, E. [Department of Astronomy and Space Science, Universita di Firenze, L.go E. Fermi 2, 50125 Firenze (Italy); INFN, Sezione di Firenze, v. G. Sansone 1, Sesto Fiorentino, Firenze (Italy); Cinque, G.; Marcelli, A. [Laboratori Nazionali di Frascati, INFN, 00044 Frascati, Roma (Italy); Achard, J.; Tallaire, A. [LIMHP-CNRS, University of Paris XIII, 99 Avenue JB Clement, 93430 Villetaneuse (France)

    2007-07-15

    Due to its unique physical properties, diamond is a very appealing material for the development of electronic devices and sensors. Its wide band gap (5.5 eV) endows diamond based devices with low thermal noise, low dark current levels and, in the case of radiation detectors, high visible-to-X-ray signal discrimination (visible blindness) as well as high sensitivity to energies greater than the band gap. Furthermore, due to its radiation hardness diamond is very interesting for applications in extreme environments, or as monitor of high fluency radiation beams. In this work the use of diamond based detectors for X-ray sensing is discussed. On purpose, some photo-conductors based on different diamond types have been tested at the DAFNE-L synchrotron radiation laboratory at Frascati. X-ray sensitivity spectra, linearity and stability of the response of these diamond devices have been measured in order to evidence the promising performance of such devices.

  8. Diamond detectors for synchrotron radiation X-ray applications

    International Nuclear Information System (INIS)

    De Sio, A.; Pace, E.; Cinque, G.; Marcelli, A.; Achard, J.; Tallaire, A.

    2007-01-01

    Due to its unique physical properties, diamond is a very appealing material for the development of electronic devices and sensors. Its wide band gap (5.5 eV) endows diamond based devices with low thermal noise, low dark current levels and, in the case of radiation detectors, high visible-to-X-ray signal discrimination (visible blindness) as well as high sensitivity to energies greater than the band gap. Furthermore, due to its radiation hardness diamond is very interesting for applications in extreme environments, or as monitor of high fluency radiation beams. In this work the use of diamond based detectors for X-ray sensing is discussed. On purpose, some photo-conductors based on different diamond types have been tested at the DAFNE-L synchrotron radiation laboratory at Frascati. X-ray sensitivity spectra, linearity and stability of the response of these diamond devices have been measured in order to evidence the promising performance of such devices

  9. X-ray Peltier cooled detectors for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Loupilov, A.; Sokolov, A.; Gostilo, V.

    2001-01-01

    The recent results on development of X-ray Si(Li), Si-planar and CdTe p-i-n detectors cooled by Peltier coolers for fabrication of laboratory and portable XRF analysers for different applications are discussed. Low detection limits of XRF analysers are provided by increasing of detectors sensitive surface; improvement of their spectrometrical characteristics; decreasing of front-end-electronics noise level; Peltier coolers and vacuum chambers cooling modes optimization. Solution of all mentioned tasks allowed to develop Peltier cooled detectors with the following performances: (1.) Si(Li) detectors: S=20 mm 2 , thickness=3.5 mm, 175 eV (5.9 keV), 430 eV (59.6 keV); S=100 mm 2 ; thickness=4.5 mm, 270 eV (5.9 keV), 485 eV (59.6 keV). (2.) Si-planar detector: S=10 mm 2 , thickness=0.4 mm, 230 eV (5.9 keV), 460 eV (59.6 keV). (3.) CdTe p-i-n detectors: S=16 mm 2 , thickness=0.5 mm, 350 eV (5.9 keV), 585 eV (59.6 keV). S=16 mm 2 , thickness=1.2 mm, 310 eV (5.9 keV), 600 eV (59.6 keV). Advantages and disadvantages of all types of detectors for X-ray fluorescence analysis are compared. Spectra are presented. Application of different XRF analysers based on developed detectors in medicine, environmental science, industry, cryminalistics and history of art are demonstrated

  10. X-ray Peltier cooled detectors for X-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Loupilov, A.; Sokolov, A.; Gostilo, V.

    2000-01-01

    The recent results on development of X-ray Si(Li), Si-planar and CdTe p-i- n detectors cooled by Peltier coolers for fabrication of laboratory and portable XRF analysers for different applications are discussed. Low detection limits of XRF analysers are provided by increasing of detectors sensitive surface; improvement of their spectrometrical characteristics; decreasing of front-end-electronics noise level; Peltier coolers and vacuum chambers cooling modes optimization. Solution of all mentioned tasks allowed to develop Peltier cooled detectors with the following performances: (1) Si(Li) detectors: S = 20 mm 2 , thickness = 3.5 mm, 175 eV (5.9 keV), 430 eV (59.6 keV); S = 100 mm 2 ; thickness = 4.5 mm, 270 eV (5.9 keV), 485 eV (59,6 keV). (2) Si-planar detector: S = 10 mm 2 , thickness = 0.4 mm, 230 eV (5.9 keV), 460 eV (59.6 keV). (3) CdTe p-i-n detectors: S = 16 mm 2 , thickness 0.5 mm, 350 eV (5.9 keV), 585 eV (59.6 keV). S = 16 mm 2 , thickness = 1.2 mm, 310 eV (5.9 keV), 600 eV (59.6 keV). Advantages and disadvantages of all types of detectors for X-ray fluorescence analysis are compared. Spectra are presented. Application of different XRF analysers based on developed detectors in medicine, environmental science, industry, criminalistics and history of art are demonstrated. (author)

  11. Gas microstrip detectors for X-ray tomographic flow imaging

    CERN Document Server

    Key, M J; Luggar, R D; Kundu, A

    2003-01-01

    A investigation into the suitability of gas microstrip detector technology for a high-speed industrial X-ray tomography system is reported. X-ray energies in the region 20-30 keV are well suited to the application, which involves imaging two-dimensional slices through gas/liquid multiphase pipeline flows for quantitative component fraction measurement. Stable operation over a period representing several hundred individual tomographic scans at gas gains of 500 is demonstrated using a Penning gas mixture of krypton/propylene.

  12. Charge diffusion in CCD X-ray detectors

    International Nuclear Information System (INIS)

    Pavlov, George G.; Nousek, John A.

    1999-01-01

    Critical to the detection of X-rays by CCDs, is the detailed process of charge diffusion and drift within the device. We reexamine the prescriptions currently used in the modeling of X-ray CCD detectors to provide analytic expressions for the charge distribution over the CCD pixels which are suitable for use in numerical simulations of CCD response. Our treatment results in models which predict charge distributions which are more centrally peaked and have flatter wings than the Gaussian shapes predicted by previous work and adopted in current CCD modeling codes

  13. Effective and cheap X-ray television detector

    International Nuclear Information System (INIS)

    Artem'ev, A.N.; Potlovskij, K.G.; Rezvov, V.A.; Yudin, L.I.

    2002-01-01

    The position sensitive detector (PSD) is designed for investigations with traditional X-ray tubes and synchrotron radiation from 3 to 30 keV. PSD consists of light-tight box, which transforms X-ray photons to light photons. Light photons are registered with the help of TV camera. Then an image is digitized and introduced into computer. Software provides registration of the dim beam images by means of accumulation of the information. Statistic processing of the image series allows to determine of the parameters of the image. Sensitivity is 41 phot/pixel. Spatial resolution is not worse then 400 μ [ru

  14. The pin pixel detector--X-ray imaging

    CERN Document Server

    Bateman, J E; Derbyshire, G E; Duxbury, D M; Marsh, A S; Simmons, J E; Stephenson, R

    2002-01-01

    The development and testing of a soft X-ray gas pixel detector, which uses connector pins for the anodes is reported. Based on a commercial 100 pin connector block, a prototype detector of aperture 25.4 mm centre dot 25.4 mm can be economically fabricated. The individual pin anodes all show the expected characteristics of small gas detectors capable of counting rates reaching 1 MHz per pin. A 2-dimensional resistive divide readout system has been developed to permit the imaging properties of the detector to be explored in advance of true pixel readout electronics.

  15. Portable X-Ray, K-Edge Heavy Metal Detector

    International Nuclear Information System (INIS)

    Fricke, V.

    1999-01-01

    The X-Ray, K-Edge Heavy Metal Detection System was designed and built by Ames Laboratory and the Center for Nondestructive Evaluation at Iowa State University. The system uses a C-frame inspection head with an X-ray tube mounted on one side of the frame and an imaging unit and a high purity germanium detector on the other side. the inspection head is portable and can be easily positioned around ventilation ducts and pipes up to 36 inches in diameter. Wide angle and narrow beam X-ray shots are used to identify the type of holdup material and the amount of the contaminant. Precise assay data can be obtained within minutes of the interrogation. A profile of the containerized holdup material and a permanent record of the measurement are immediately available

  16. Charge-coupled-device X-ray detector performance model

    Science.gov (United States)

    Bautz, M. W.; Berman, G. E.; Doty, J. P.; Ricker, G. R.

    1987-01-01

    A model that predicts the performance characteristics of CCD detectors being developed for use in X-ray imaging is presented. The model accounts for the interactions of both X-rays and charged particles with the CCD and simulates the transport and loss of charge in the detector. Predicted performance parameters include detective and net quantum efficiencies, split-event probability, and a parameter characterizing the effective thickness presented by the detector to cosmic-ray protons. The predicted performance of two CCDs of different epitaxial layer thicknesses is compared. The model predicts that in each device incomplete recovery of the charge liberated by a photon of energy between 0.1 and 10 keV is very likely to be accompanied by charge splitting between adjacent pixels. The implications of the model predictions for CCD data processing algorithms are briefly discussed.

  17. Gaseous detectors for energy dispersive X-ray fluorescence analysis

    Science.gov (United States)

    Veloso, J. F. C. A.; Silva, A. L. M.

    2018-01-01

    The energy resolution capability of gaseous detectors is being used in the last years to perform studies on the detection of characteristic X-ray lines emitted by elements when excited by external radiation sources. One of the most successful techniques is the Energy Dispersive X-ray Fluorescence (EDXRF) analysis. Recent developments in the new generation of micropatterned gaseous detectors (MPGDs), triggered the possibility not only of recording the photon energy, but also of providing position information, extending their application to EDXRF imaging. The relevant features and strategies to be applied in gaseous detectors in order to better fit the requirements for EDXRF imaging will be reviewed and discussed, and some application examples will be presented.

  18. Recent developments in X-ray imaging detectors

    CERN Document Server

    Moy, J P

    2000-01-01

    The replacement of the radiographic film in medical imaging has been the driving force in X-ray imaging developments. It requires a approx 40 cm wide detector to cover all examinations, an equivalent noise level of 1-5 X-ray quanta per pixel, and spatial resolution in the range 100-150 mu m. The need for entirely electronic imaging equipments has fostered the development of many X-ray detectors, most of them based on an array of amorphous silicon pixels, which is the only technology capable to achieve such large areas. Essentially, two concepts have been implemented: - intermediate conversion of X-rays to light by a scintillator, detected by an array of light sensitive pixels, comprising a photodiode and a switching device, either a TFT or a diode. - conversion into electron-hole pairs in a photoconductor, collected by an array of electrodes and switches. In both cases, charge amplifiers read the generated charges line by line. Scintillator and photoconductor-based systems are now close to production. They ac...

  19. Developments in gas detectors for synchrotron x-ray radiation

    International Nuclear Information System (INIS)

    Fischer, J.; Radeka, V.; Smith, G.C.

    1985-09-01

    New results on the physical limitations to position resolution in gas detectors for x-rays (approx. =3 to 20 keV) due to the range of photoelectrons and Auger electrons are discussed. These results were obtained with a small gap detector in which position readout was accomplished by using a very low noise centroid finding technique. A description is given of position sensitive detectors for medium rates (a few x 10 5 photons per second), using delay line readout, and for very high rates (approx. =10 8 photons per second), using fast signal shaping on the output of each anode wire

  20. Assessing the efficiency position sensitive gaseous X-rays detectors

    International Nuclear Information System (INIS)

    Almeida, Gevaldo L. de; Souza, Maria Ines Silvani; Lopes, Ricardo T.

    2009-01-01

    Full text: The efficiency of gaseous X-ray detectors can be evaluated from tabulated data, but this approach assumes that the whole detector volume is permeated by the electrostatic field produced by the anode-cathode. Indeed, the usual detectors are comprised by a cylindrical hull acting as cathode containing a wire at its axis as anode, a configuration which foods the space between them with the electrostatic field. Some specially designed detectors, however, as Position Sensitive Detectors, contain regions which are not submitted to the electrostatic field, and hence, their efficiency could not be assessed from the tabulated data. Direct measurements of this efficiency would require a mono-chromator or set of pure mono-energetic X-rays sources. As only very few of them are really mono-energetic, the detector response to a given energy would be spoiled by to the concomitant contribution of other energies. Yet, the information would not be completely lost, but only concealed due to the convolution carried out by the detector. Therefore, a proper unfolding would be capable to recover the information, yielding the individual detector efficiency for each of the contributing energies. The degraded information is retrieved in this work through a proper mathematical unfolding of the detector response, when exposed to Bremsstrahlung spectra from an X-ray tube submitted to different voltages. For this purpose, Lorentzian functions have been fitted to these spectra - obtained with a NaI(Tl) spectrometer - in order to characterize them with proper parameters. The mathematical convolution of these functions with a theoretical detector efficiency curve yields, after integration, values which, confronted with those experimentally measured, allow the determination of the parameters of the efficiency curve. As some parameters of this curve are well known, it is possible to represent it by proper functions. For argon-filled detectors, for instance, this efficiency has a

  1. Gas pixel detector for X-ray observation

    CERN Document Server

    Attié, D; Chefdeville, M; Colas, P; Delagnes, E; Giomataris, Y; van der Graaf, H; Llopart, X; Timmermans, J; Visschers, J

    2009-01-01

    We report on the status of the R&D for a digital Time Projection Chamber (TPC) based on Micromegas (MICRO MEsh GAseous Structure) detectors using the CMOS chip TimePix as a direct readout anode protected by highly resistive a-Si:H layer. A small chamber was built as a demonstrator of the 2-D and 3-D imaging capabilities of this technique. We illustrate the new capabilities of this detector for X-ray observation with data taken from radioactive sources. This small TPC is a very useful tool both for studying gas properties thanks to its good efficiency for single electrons, and for reconstructing photoelectron direction for use as a soft X-ray polarimeter.

  2. Position sensitive detector for X-ray photons

    International Nuclear Information System (INIS)

    Barbosa, A.F.

    1988-01-01

    This work reports the theoretical basis and the details of the construction process, characterization and application of gas X-ray position sensitive detectors. The unidimensional detector consists of a gas camera (argon and CH 4 ), a metallic anode, a cathode and a delay line. Details of the construction process are given in order to allow the reproduction of the detector. It has been characterized by measuring its spatial resolution, homogeneity and linerity. The built linear detector has been used to obtain diffraction diagrams from polycrystalline silicon, C 23 H 48 paraffin and glassy carbon. These diagrams have been compared with those obtained under equivalent conditions with a conventional proportional detector by the step scanning method. It has been shown that the detector provides diffraction diagrams of equivalent quality to those obtained by the step scanning method, in appreciably lower time intervals. (author) [pt

  3. The color of X-rays Spectral X-ray computed tomography using energy sensitive pixel detectors

    CERN Document Server

    Schioppa, Enrico Junior

    Energy sensitive X-ray imaging detectors are produced by connecting a semiconductor sensor to a spectroscopic pixel readout chip. In this thesis, the applicability of such detectors to X-ray Computed Tomography (CT) is studied. A prototype Medipix based silicon detector is calibrated using X-ray fluorescence. The charge transport properties of the sensor are characterized using a high energy beam of charged particles at the Super Proton Synchrotron (SPS) at the European Center for Nuclear Research (CERN). Monochromatic X-rays at the European Synchrotron Radiation Facility (ESRF) are used to determined the energy response function. These data are used to implement a physics-based CT projection operator that accounts for the transmission of the source spectrum through the sample and detector effects. Based on this projection operator, an iterative spectral CT reconstruction algorithm is developed by extending an Ordered Subset Expectation Maximization (OSEM) method. Subsequently, a maximum likelihood based algo...

  4. X-ray measurement with Pin type semiconductor detectors

    International Nuclear Information System (INIS)

    Ramirez J, F.J.

    1999-01-01

    Here are presented the experimental results of the applications of Pin type radiation detectors developed in a National Institute of Nuclear Research (ININ) project, in the measurement of low energy gamma and X-rays. The applications were oriented mainly toward the Medical Physics area. It is planned other applications which are in process of implementation inside the National Institute of Nuclear Research in Mexico. (Author)

  5. Scintillator materials for x-ray detectors and beam monitors

    Czech Academy of Sciences Publication Activity Database

    Martin, T.; Koch, A.; Nikl, Martin

    2017-01-01

    Roč. 42, č. 6 (2017), s. 451-456 ISSN 0883-7694 R&D Projects: GA ČR GA16-15569S Institutional support: RVO:68378271 Keywords : scintillator * X-ray detector * beam monitor * synchrotron * thin film Subject RIV: BM - Solid Matter Physics ; Magnetism OBOR OECD: Condensed matter physics (including formerly solid state physics, supercond.) Impact factor: 5.199, year: 2016

  6. Large area x-ray detectors for cargo radiography

    Science.gov (United States)

    Bueno, C.; Albagli, D.; Bendahan, J.; Castleberry, D.; Gordon, C.; Hopkins, F.; Ross, W.

    2007-04-01

    Large area x-ray detectors based on phosphors coupled to flat panel amorphous silicon diode technology offer significant advances for cargo radiologic imaging. Flat panel area detectors provide large object coverage offering high throughput inspections to meet the high flow rate of container commerce. These detectors provide excellent spatial resolution when needed, and enhanced SNR through low noise electronics. If the resolution is reduced through pixel binning, further advances in SNR are achievable. Extended exposure imaging and frame averaging enables improved x-ray penetration of ultra-thick objects, or "select-your-own" contrast sensitivity at a rate many times faster than LDAs. The areal coverage of flat panel technology provides inherent volumetric imaging with the appropriate scanning methods. Flat panel area detectors have flexible designs in terms of electronic control, scintillator selection, pixel pitch, and frame rates. Their cost is becoming more competitive as production ramps up for the healthcare, nondestructive testing (NDT), and homeland protection industries. Typically used medical and industrial polycrystalline phosphor materials such as Gd2O2S:Tb (GOS) can be applied to megavolt applications if the phosphor layer is sufficiently thick to enhance x-ray absorption, and if a metal radiator is used to augment the quantum detection efficiency and reduce x-ray scatter. Phosphor layers ranging from 0.2-mm to 1-mm can be "sandwiched" between amorphous silicon flat panel diode arrays and metal radiators. Metal plates consisting of W, Pb or Cu, with thicknesses ranging from 0.25-mm to well over 1-mm can be used by covering the entire area of the phosphor plate. In some combinations of high density metal and phosphor layers, the metal plate provides an intensification of 25% in signal due to electron emission from the plate and subsequent excitation within the phosphor material. This further improves the SNR of the system.

  7. An x-ray detector using superconducting aluminum tunnel junctions

    International Nuclear Information System (INIS)

    Barber, W.C.; Bland, R.W.; Carpenter, J.W.; Johnson, R.T.; Laws, K.E.; Lockhart, J.; Lee, J.S.; Watson, R.M.

    1992-01-01

    We report on tests of a prototype detector for 6-keV X-rays, using series arrays of tunnel junction. Tests with higher-energy particles indicate an energy resolution of 4 keV, at 0.3K and with a warm pre-amp. At lower temperatures and with a cooled FET, the resolution should approach 100 eV

  8. An X-ray gas position sensitive detector: construction and characterization

    International Nuclear Information System (INIS)

    Barbosa, A.F.; Gabriel, A.; Gabriel, A.; Craievich, A.

    1988-01-01

    A linear x-ray gas position sensitive detector with delay line readout has been constructed. The detector is described, characterized and used for detecting x-ray diffraction patterns from polycrystals. (author) [pt

  9. Flat-response x-ray-diode-detector development

    International Nuclear Information System (INIS)

    Tirsell, G.

    1982-10-01

    In this report we discuss the design of an improved sub-nanosecond flat response x-ray diode detector needed for ICF diagnostics. This device consists of a high Z cathode and a complex filter tailored to flatten the response so that the total x-ray energy below 1.5 keV can be measured using a single detector. Three major problems have become evident as a result of our work with the original LLNL design including deviation from flatness due to a peak in the response below 200 eV, saturation at relatively low x-ray fluences, and long term gold cathode instability. We are investigating grazing incidence reflection to reduce the response below 200 eV, new high Z cathode materials for long term stability, and a new complex filter for improved flatness. Better saturation performance will require a modified XRD detector under development with reduced anode to cathode spacing and increased anode bias voltage

  10. Quality control measurements for digital x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Marshall, N W [Department of Radiology, University Hospitals Leuven, 49 Herenstraat, 3000 Leuven (Belgium); Mackenzie, A [National Co-ordinating Centre for the Physics of Mammography, Medical Physics, Level B, St Luke' s Wing, The Royal Surrey County Hospital NHS Trust, Egerton Road, Guildford, GU2 7XX (United Kingdom); Honey, I D, E-mail: nicholas.marshall@uz.kuleuven.ac.be [Department of Medical Physics, Floor 3, Henriette Raphael House, Guy' s and St Thomas' Hospital, London, SE1 9RT (United Kingdom)

    2011-02-21

    This paper describes a digital radiography (DR) quality control protocol for DR detectors from the forthcoming report from the Institute of Physics and Engineering in Medicine (IPEM). The protocol was applied to a group of six identical caesium iodide (CsI) digital x-ray detectors to assess reproducibility of methods, while four further detectors were assessed to examine the wider applicability. Twelve images with minimal spatial frequency processing are required, from which the detector response, lag, modulation transfer function (MTF), normalized noise power spectrum (NNPS) and threshold contrast-detail (c-d) detectability are calculated. The x-ray spectrum used was 70 kV and 1 mm added copper filtration, with a target detector air kerma of 2.5 {mu}Gy for the NNPS and c-d results. In order to compare detector performance with previous imaging technology, c-d data from four screen/film systems were also acquired, at a target optical density of 1.5 and an average detector air kerma of 2.56 {mu}Gy. The DR detector images were typically acquired in 20 min, with a further 45 min required for image transfer and analysis. The average spatial frequency for the 50% point of the MTF for six identical detectors was 1.29 mm{sup -1} {+-} 0.05 (3.9% coefficient of variation (cov)). The air kerma set for the six systems was 2.57 {mu}Gy {+-} 0.13 (5.0% cov) and the NNPS at this air kerma was 1.42 x 10{sup -5} mm{sup 2} (6.5% cov). The detective quantum efficiency (DQE) measured for the six identical detectors was 0.60 at 0.5 mm{sup -1}, with a maximum cov of 10% at 2.9 mm{sup -1}, while the average DQE was 0.56 at 0.5 mm{sup -1} for three CsI detectors from three different manufacturers. Comparable c-d performance was found for these detectors (5.9% cov) with an average threshold contrast of 0.46% for 11 mm circular discs. The average threshold contrast for the S/F systems was 0.70% at 11 mm, indicating superior imaging performance for the digital systems. The protocol was found

  11. Time response characteristics of X-ray detector system on Silex-Ⅰ laser facility

    International Nuclear Information System (INIS)

    Yi Rongqing; He Xiao'an; Li Hang; Du Huabing; Zhang Haiying; Cao Zhurong

    2013-01-01

    On the Silex-Ⅰ laser facility, the time response characteristics of XRD detector were studied. A laser with a pulse of 32 fs and a wavelength of 800 nm was used to irradiate a plane Au target. X-ray calibrated method of time of exposure X-ray framing camera and time resolution of X-ray streak camera was explored. The time response characteristics of XRD detector and time process of X-ray emission were obtained from experiment. We obtained X-ray calibration method of time of exposure X-ray framing camera and time resolution of X-ray streak camera. (authors)

  12. Self-propelled x-ray flaw detector

    International Nuclear Information System (INIS)

    Ershov, L.S.; Krasilnikov, S.B.; Lozovoi, L.N.; Losev, J.F.; Morgovsky, L.Y.; Pelix, E.A.; Khakimyanov, R.R.

    1988-01-01

    A self-propelled X-ray flaw detector for radiographic inspection of welded joints in pipelines comprises a carriage mounting a motor, a detector having two Geiger counters, a pulsed X-ray generator, and an exposure and carriage electronic control system. A memory unit in the control system has four storage elements containing information about the motion of the carriage. As the carriage moves in direction A, first one and then the other of the Geiger counters receives radiation from an isotope source positioned near a joint, and by means of logic circuitry in the control system, the information in the storage elements is modified to stop the carriage and to operate a timer to expose the weld. During exposure the X-rays may interfere with the information in the storage elements, but by means of a square-wave generator and the logic circuitry, the stored information is correctly reset in order to eliminate false operation of the memory unit. (author)

  13. Diagnostic X-ray spectrometry using a commercial CdZnTe detector

    International Nuclear Information System (INIS)

    Becker, P.H.B.

    1998-01-01

    X ray spectrometry using Ge or Si detectors is an established tool to measure characterization parameters of X-ray beams. This work describes how a commercial CdZnTe was used to perform diagnostic X-ray spectrometry. Spectra were measured for two X-ray machines and compared with similar data found in the literature with an agreement of 2% rms

  14. Detectors for X-ray diffraction and scattering: a user's overview

    International Nuclear Information System (INIS)

    Bruegemann, Lutz; Gerndt, E.K.E.

    2004-01-01

    An overview of the applications of X-ray detectors to material research is given. Four experimental techniques and their specific detector requirements are described. Detector types are classified and critical parameters described in the framework of X-ray diffraction and X-ray scattering experiments. The article aims at building a bridge between detector end-users and detector developers. It gives limits of critical detector parameters, like angular resolution, energy resolution, dynamic range, and active area

  15. Radiation damage resistance in mercuric iodide X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Patt, B E; Dolin, R C; Devore, T M; Markakis, J M [EG and G Energy Measurements, Inc., Goleta, CA (USA); Iwanczyk, J S; Dorri, N [Xsirius, Inc., Marina del Rey, CA (USA); Trombka, J [National Aeronautics and Space Administration, Greenbelt, MD (USA). Goddard Space Flight Center

    1990-12-20

    Mercuric iodide (HgI{sub 2}) radiation detectors show great potential as ambient-temperature solid-state detectors for X-rays, gamma rays and visible light, with parameters that are competitive with existing technologies. In a previous experiment, HgI{sub 2} detectors irradiated with 10 MeV protons/cm{sup 2} exhibited no damage. The 10 MeV protons represent only the low range of the spectrum of energies that are important. An experiment has been conducted at the Saturne accelerator facility at Saclay, France, to determine the susceptibility of these detectors to radiation damage by high-energy (1.5 GeV) protons. The detectors were irradiated to a fluence of 10{sup 8} protons/cm{sup 2}. This fluence is equivalent to the cosmic radiation expected in a one-year period in space. The resolution of the detectors was measured as a function of the integral dose. No degradation in the response of any of the detectors or spectrometers was seen. It is clear from this data that HgI{sub 2} has extremely high radiation-damage resistance, exceeding that of most other semiconductor materials used for radiation detectors. Based on the results shown to date, HgI{sub 2} detectors are suitable for applications in which they may be exposed to high integral dose levels. (orig.).

  16. Calculation of Si(Li) x-ray detector efficiencies

    International Nuclear Information System (INIS)

    Zaluzec, N.; Holton, R.

    1984-01-01

    The calculation of detector efficiency functions is an important step in the quantitative analysis of x-ray spectra when approached by a standardless technique. In this regard, it becomes essential that the analyst not only model the physical aspects of the absorption and transmission of the various windows present, but also use the most accurate data available for the mass absorption coefficients required in these calculations. The topic of modeling the size and shape of the windows present is beyond the scope of this paper and the authors instead concentrate on the mass absorption coefficients used in the calculations and their implications to efficiency calculations. For the purposes of this paper, the authors consider that the relative detector efficiency function of a conventional Si(Li) detector can be modeled by a simple expression

  17. CONTINUING THE DEVELOPMENT OF A 100 FEMTOSECOND X-RAY DETECTOR

    International Nuclear Information System (INIS)

    Zenghu Chang

    2005-01-01

    The detector is an x-ray streak camera running in accumulation mode for time resolved x-ray studies at the existing third generation synchrotron facilities and will also be used for the development and applications of the fourth generation x-ray sources. We have made significant progress on both the detector development and its applications at Synchrotron facilities

  18. Pixelated transmission-mode diamond X-ray detector.

    Science.gov (United States)

    Zhou, Tianyi; Ding, Wenxiang; Gaowei, Mengjia; De Geronimo, Gianluigi; Bohon, Jen; Smedley, John; Muller, Erik

    2015-11-01

    Fabrication and testing of a prototype transmission-mode pixelated diamond X-ray detector (pitch size 60-100 µm), designed to simultaneously measure the flux, position and morphology of an X-ray beam in real time, are described. The pixel density is achieved by lithographically patterning vertical stripes on the front and horizontal stripes on the back of an electronic-grade chemical vapor deposition single-crystal diamond. The bias is rotated through the back horizontal stripes and the current is read out on the front vertical stripes at a rate of ∼ 1 kHz, which leads to an image sampling rate of ∼ 30 Hz. This novel signal readout scheme was tested at beamline X28C at the National Synchrotron Light Source (white beam, 5-15 keV) and at beamline G3 at the Cornell High Energy Synchrotron Source (monochromatic beam, 11.3 keV) with incident beam flux ranges from 1.8 × 10(-2) to 90 W mm(-2). Test results show that the novel detector provides precise beam position (positional noise within 1%) and morphology information (error within 2%), with an additional software-controlled single channel mode providing accurate flux measurement (fluctuation within 1%).

  19. Comparison of natural and synthetic diamond X-ray detectors

    International Nuclear Information System (INIS)

    Lansley, S. P.; Betzel, G.T.; Meyer, J.; Metcalf, P.; Reinisch, L.

    2010-01-01

    , the synthetic diamond detector performed well in comparison to the natural dia mond detector. Keywords X-ray detector Synthetic diamond Natural diamond Energy dependence Dose linearity . Dose-rate dependence X-ray detectors fabricated from diamond have been co sidered for many years, e.g. [1-5], for reasons including the near-tissue equivalence and high radiation tolerance of diamond, and low leakage current (which helps improve signal-noise ratio) due to high intrinsic resistivity. It is worth noting that, while diamond is described as near-tis sue equivalent because of its atomic number (Z = 6), the mass density of diamond (3.51 g/cm3) is much higher than water (1.00 g/cm/ muscle (1.06 g/cm3) [6] or fat (0.92 gl cm3) [7]. Natural diamond-based detectors for use in high-energy photon and electron beams are only commercially-avai able from Physikalisch- Technische Werkstatten GmbH (PTW, Freiburg, Germany) [8]; they were developed in cooperation with the IPTP Institute in Riga, Latvia. These detectors have a small sensitive volume (1-6 mm3) and hence are marketed as being especially well suited for

  20. Radio and X-ray emission from newly born remnants

    International Nuclear Information System (INIS)

    Salvati, M.

    1983-01-01

    Radio and X-ray observations of SN 1979c and SN 1980k offer a unique opportunity of monitoring the transition from supernovae to remnants. By means of the two-frequency radio light curves, the hypothesis that these objects are surrounded by circumstellar matter, originated in a presupernova wind, is tested, and the relevant parameters are derived. Then the absorption-corrected light curves are used to test the various proposed models. SN 1980k appears to be powered by a canonical shock, while SN 1979c is a good plerion candidate. An optical pulsar could still be detected at its location. (Auth.)

  1. X-ray Emission from Solar Flares

    Indian Academy of Sciences (India)

    2016-01-27

    Jan 27, 2016 ... Solar flares; X-ray detectors; X-ray line emission and continuum; break energy; microflares. Abstract. Solar X-ray Spectrometer (SOXS), the first space-borne solar astronomy experiment of India was designed to improve our current understanding of X-ray emission from the Sun in general and solar flares in ...

  2. X-ray detectors at the Linac Coherent Light Source

    International Nuclear Information System (INIS)

    Blaj, Gabriel; Caragiulo, Pietro; Carini, Gabriella; Carron, Sebastian; Dragone, Angelo; Freytag, Dietrich; Haller, Gunther; Hart, Philip; Hasi, Jasmine; Herbst, Ryan; Herrmann, Sven; Kenney, Chris; Markovic, Bojan; Nishimura, Kurtis; Osier, Shawn; Pines, Jack; Reese, Benjamin; Segal, Julie; Tomada, Astrid; Weaver, Matt

    2015-01-01

    This paper offers an overview of area detectors developed for use at the Linac Coherent Light Source (LCLS) with particular emphasis on their impact on science. The experimental needs leading to the development of second-generation cameras for LCLS are discussed and the new detector prototypes are presented. Free-electron lasers (FELs) present new challenges for camera development compared with conventional light sources. At SLAC a variety of technologies are being used to match the demands of the Linac Coherent Light Source (LCLS) and to support a wide range of scientific applications. In this paper an overview of X-ray detector design requirements at FELs is presented and the various cameras in use at SLAC are described for the benefit of users planning experiments or analysts looking at data. Features and operation of the CSPAD camera, which is currently deployed at LCLS, are discussed, and the ePix family, a new generation of cameras under development at SLAC, is introduced

  3. Synchrotron radiation calibration for soft X-ray detector

    International Nuclear Information System (INIS)

    Ning, Jiamin; Guo, Cun; Xu, Rongkun; Jiang, Shilun; Xu, Zeping; Chen, Jinchuan; Xia, Guangxin; Xue, Feibiao; Qin, Yi

    2009-04-01

    The calibration experiments were carried out to X-ray film, scintillator and transmission grating by employing the soft X-ray station at 3W1B beam-line in Beijing synchrotron Radiation Facility. The experiments presented the black intensity curve and energy response curve of soft X-ray film. And the experimental results can be used in diagnosis of X-ray radiation characterization of Z-pinch, such as in the measurement of soft X-ray Power Meter, grating spectrometer, pinhole camera and one-dimension imaging system which can ensure precision of Z-pinch results. (authors)

  4. X-ray detector for automatic exposure control using ionization chamber filled with xenon gas

    CERN Document Server

    Nakagawa, A; Yoshida, T

    2003-01-01

    This report refers to our newly developed X-ray detector for reliable automatic X-ray exposure control, which is to be widely used for X-ray diagnoses in various clinical fields. This new detector utilizes an ionization chamber filled with xenon gas, in contrast to conventional X-ray detectors which use ionization chambers filled with air. Use of xenon gas ensures higher sensitivity and thinner design of the detector. The xenon gas is completely sealed in the chamber, so that the influence of the changes in ambient environments is minimized. (author)

  5. Photoemission measurements for low energy x-ray detector applications

    International Nuclear Information System (INIS)

    Day, R.H.

    1981-01-01

    Photoemission has been studied for nearly 100 years as both a means of investigating quantum physics, and as a practical technique for transducing optical/x-ray photons into electrical currents. Numerous x-ray detection schemes, such as streak cameras and x-ray sensitive diodes, exploit this process because of its simplicity, adaptability, and speed. Recent emphasis on diagnostics for low temperature, high density, and short-lived, plasmas for inertial confinement fusion has stimulated interest in x-ray photoemission in the sub-kilovolt regime. In this paper, a review of x-ray photoemission measurements in the 50 eV to 10 keV x-ray region is given and the experimental techniques are reviewed. A semiempirical model of x-ray photoemission is discussed and compared to experimental measurements. Finally, examples of absolutely calibrated instruments are shown

  6. Novel X-ray Communication Based XNAV Augmentation Method Using X-ray Detectors

    Directory of Open Access Journals (Sweden)

    Shibin Song

    2015-09-01

    Full Text Available The further development of X-ray pulsar-based NAVigation (XNAV is hindered by its lack of accuracy, so accuracy improvement has become a critical issue for XNAV. In this paper, an XNAV augmentation method which utilizes both pulsar observation and X-ray ranging observation for navigation filtering is proposed to deal with this issue. As a newly emerged concept, X-ray communication (XCOM shows great potential in space exploration. X-ray ranging, derived from XCOM, could achieve high accuracy in range measurement, which could provide accurate information for XNAV. For the proposed method, the measurement models of pulsar observation and range measurement observation are established, and a Kalman filtering algorithm based on the observations and orbit dynamics is proposed to estimate the position and velocity of a spacecraft. A performance comparison of the proposed method with the traditional pulsar observation method is conducted by numerical experiments. Besides, the parameters that influence the performance of the proposed method, such as the pulsar observation time, the SNR of the ranging signal, etc., are analyzed and evaluated by numerical experiments.

  7. Imaging plate, a new type of x-ray area detector

    International Nuclear Information System (INIS)

    Kamiya, Nobuo; Amemiya, Yoshiyuki; Miyahara, Junji.

    1986-01-01

    In respective fields of X-ray crystallography, for the purpose of the efficient collection of reciprocal space information, two-dimensional X-ray detectors such as multiwire proportional chambers and X-ray television sets have been used together with conventional X-ray films. X-ray films are characterized by uniform sensitivity and high positional resolution over a wide area, but the sensitivity is low, and the range of action and the linearity of the sensitivity is problematic. They require the development process, accordingly lack promptitude. The MWPCs and X-ray television sets are superior in the sensitivity, its linearity, the range of action and promptitude, but interior in the uniformity and resolution to the films. Imaging plate is a new X-ray area detector developed by Fuji Photo Film Co., Ltd., for digital X-ray medical image diagnosis. This detector is superior in all the above mentioned performances, and it seems very useful also for X-ray crystallography. In this paper, the system composed of an imaging plate and its reader is described, and the basic performance as an X-ray area detector and the results of having recorded the diffraction images of protein crystals as the example of applying it to X-ray crystallography are reported. The imaging plate is that the crystalline fluorescent powder of BaFBr doped with Eu 2+ ions is applied on plastic films. (Kako, I.)

  8. A large area detector for x-ray applications

    International Nuclear Information System (INIS)

    Rodricks, B.; Huang, Qiang; Hopf, R.; Wang, Kemei.

    1993-01-01

    A large area detector for x-ray synchrotron applications has been developed. The front end of this device consist of a scintillator coupled to a fiber-optic taper. The fiber-optic taper is comprised of 4 smaller (70 mm x 70 mm) tapers fused together in a square matrix giving an active area of 140 mm x 140 mm. Each taper has a demagnification of 5.5 resulting in four small ends that are 12 mm diagonally across. The small ends of each taper are coupled to four microchannel-plate-based image intensifiers. The output from each image intensifier is focused onto a Charge Coupled Device (CCD) detector. The four CCDs are read out in parallel and are independently controlled. The image intensifiers also act as fast (20 ns) electronic shutters. The system is capable of displaying images in real time. Additionally, with independent control on the readout of each row of data from the CCD, the system is capable of performing high speed imaging through novel readout manipulation

  9. Radioisotope decontamination of X-ray detector. Photostimulable phosphor plate

    International Nuclear Information System (INIS)

    Onuma, Yoji; Hayashi, Michiko; Hayashi, Hiroaki; Nishihara, Sadamitsu

    2012-01-01

    We tried to remove contamination of radioisotope (RI) for an X-ray detector (photostimulable phosphor plate; IP) and verified that our procedure suggested by Nishihara et al. was effective for decontamination. The procedure was as follows. First, the IP was kept for approximately twelve hours, and then it was processed [image (A)] as well as a clinical processing mode. Second, using a wet-type chemical wiper, we scavenged the IP to remove the adhered RI on its surface. Then, once again, the IP was kept for approximately fifteen hours and processed [image (B)] in order to check an effect of decontamination. Finally, the two images of (A) and (B) were analyzed using ImageJ, which can be downloaded as a free software, and a percentage of removal was calculated. The procedure was applied to two IPs using the Fuji computed tomography (FCR) 5501 plus. In the present case, the percentage of removal was approximately 96%. The removed radioisotopes in the chemical wipers were analyzed by Ge detector. Then, 134 Cs and 137 Cs were found with activities of 2.9 4.3 Bq and 3.5 5.2 Bq, respectively. For three months after that, we cannot see black spots on the IPs owing to the contamination of the RI and there are no defects caused by decontamination using a wet-type chemical wiper. (author)

  10. Novel Hybrid CMOS X-ray Detector Developments for Future Large Area and High Resolution X-ray Astronomy Missions

    Science.gov (United States)

    Falcone, Abe

    In the coming years, X-ray astronomy will require new soft X-ray detectors that can be read very quickly with low noise and can achieve small pixel sizes over a moderately large focal plane area. These requirements will be present for a variety of X-ray missions that will attempt to address science that was highly ranked by the 2010 Decadal Survey, including missions with science that overlaps with that of IXO and Athena, as well as other missions addressing science topics beyond those of IXO and Athena. An X-ray Surveyor mission was recently chosen by NASA for study by a Science & Technology Definition Team (STDT) so it can be considered as an option for an upcom-ing flagship mission. A mission such as this was endorsed by the NASA long term planning document entitled "Enduring Quests, Daring Visions," and a detailed description of one possible reali-zation of such a mission has been referred to as SMART-X, which was described in a recent NASA RFI response. This provides an example of a future mission concept with these requirements since it has high X-ray throughput and excellent spatial resolution. We propose to continue to modify current active pixel sensor designs, in particular the hybrid CMOS detectors that we have been working with for several years, and implement new in-pixel technologies that will allow us to achieve these ambitious and realistic requirements on a timeline that will make them available to upcoming X-ray missions. This proposal is a continuation of our program that has been work-ing on these developments for the past several years. The first 3 years of the program led to the development of a new circuit design for each pixel, which has now been shown to be suitable for a larger detector array. The proposed activity for the next four years will be to incorporate this pixel design into a new design of a full detector array (2k×2k pixels with digital output) and to fabricate this full-sized device so it can be thoroughly tested and

  11. Semiconductor detectors in current energy dispersive X-ray spectral analysis

    Energy Technology Data Exchange (ETDEWEB)

    Betin, J; Zhabin, E; Krampit, I; Smirnov, V

    1980-04-01

    A review is presented of the properties of semiconductor detectors and of the possibilities stemming therefrom of using the detectors in X-ray spectral analysis in industries, in logging, in ecology and environmental control, in medicine, etc.

  12. MOS solid-state detector arrays for x-ray imaging

    International Nuclear Information System (INIS)

    Koppel, L.N.

    1977-01-01

    Two types of MOS detector arrays were used to sense directly patterns of soft x-rays, in the Lawrence Livermore Laboratory experimental laser-fusion program. A linear self-scanning photodiode array (SSPA) is used in a wave-length-dispersive spectrometer. A frame transfer charge-coupled device (CCD) facilitates the use of an x-ray microscope. Measurements and calculations of the x-ray sensitivity of these devices are presented. Their linearity and dynamic range are discussed, as well as data recovery systems for each detector. Experiences in using these devices to detect pulses of x-rays in laser-fusion experiments are described

  13. A gas microstrip wide angle X-ray detector for application in synchrotron radiation experiments

    CERN Document Server

    Bateman, J E; Derbyshire, G E; Duxbury, D M; Lipp, J; Mir, J A; Simmons, J E; Spill, E J; Stephenson, R; Dobson, B R; Farrow, R C; Helsby, W I; Mutikainen, R; Suni, I

    2002-01-01

    The Gas Microstrip Detector has counting rate capabilities several orders of magnitude higher than conventional wire proportional counters while providing the same (or better) energy resolution for X-rays. In addition the geometric flexibility provided by the lithographic process combined with the self-supporting properties of the substrate offers many exciting possibilities for X-ray detectors, particularly for the demanding experiments carried out on Synchrotron Radiation Sources. Using experience obtained in designing detectors for Particle Physics we have developed a detector for Wide Angle X-ray Scattering studies. The detector has a fan geometry which makes possible a gas detector with high detection efficiency, sub-millimetre spatial resolution and good energy resolution over a wide range of X-ray energy. The detector is described together with results of experiments carried out at the Daresbury Laboratory Synchrotron Radiation Source.

  14. Gas scintillation glass GEM detector for high-resolution X-ray imaging and CT

    Energy Technology Data Exchange (ETDEWEB)

    Fujiwara, T., E-mail: fujiwara-t@aist.go.jp [Research Institute for Measurement and Analytical Instrumentation, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Ibaraki 305-8568 (Japan); Mitsuya, Y. [Nuclear Professional School, The University of Tokyo, Tokai, Naka, Ibaraki 319-1188 (Japan); Fushie, T. [Radiment Lab. Inc., Setagaya, Tokyo 156-0044 (Japan); Murata, K.; Kawamura, A.; Koishikawa, A. [XIT Co., Naruse, Machida, Tokyo 194-0045 (Japan); Toyokawa, H. [Research Institute for Measurement and Analytical Instrumentation, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Ibaraki 305-8568 (Japan); Takahashi, H. [Institute of Engineering Innovation, School of Engineering, The University of Tokyo, Bunkyo, Tokyo 113-8654 (Japan)

    2017-04-01

    A high-spatial-resolution X-ray-imaging gaseous detector has been developed with a single high-gas-gain glass gas electron multiplier (G-GEM), scintillation gas, and optical camera. High-resolution X-ray imaging of soft elements is performed with a spatial resolution of 281 µm rms and an effective area of 100×100 mm. In addition, high-resolution X-ray 3D computed tomography (CT) is successfully demonstrated with the gaseous detector. It shows high sensitivity to low-energy X-rays, which results in high-contrast radiographs of objects containing elements with low atomic numbers. In addition, the high yield of scintillation light enables fast X-ray imaging, which is an advantage for constructing CT images with low-energy X-rays.

  15. A new X-ray detector for magnetic circular dichroism experiments

    CERN Document Server

    Bateman, J E; Dudzik, E; Laan, G V D; Lipp, J D; Smith, A D; Stephenson, R

    2001-01-01

    X-ray magnetic circular dichroism (XMCD) studies of magnetic 3d transition metal samples require the recording of high quality absorption scans in high magnetic fields using circularly polarised soft X-rays of energies in the range 0.5-1 keV. A Gas Microstrip Detector is described which permits the option of using the X-ray fluorescence signal instead of the usual electron yield signal.

  16. A new MBE CdTe photoconductor array detector for X-ray applications

    International Nuclear Information System (INIS)

    Yoo, S.S.; Sivananthan, S.; Faurie, J.P.; Rodricks, B.; Bai, J.; Montano, P.A.; Argonne National Lab., IL

    1994-10-01

    A CdTe photoconductor array x-ray detector was grown using Molecular Beam Epitaxially (MBE) on a Si (100) substrate. The temporal response of the photoconductor arrays is as fast as 21 psec risetime and 38 psec Full Width Half Maximum (FWHM). Spatial and energy responses were obtained using x-rays from a rotating anode and synchrotron radiation source. The spatial resolution of the photoconductor was good enough to provide 75 microm FWHM using a 50 microm synchrotron x-ray beam. A substantial number of x-ray photons are absorbed effectively within the MBE CdTe layer as observed from the linear response up to 15 keV. These results demonstrate that MBE grown CdTe is a suitable choice of the detector materials to meet the requirements for x-ray detectors in particular for the new high brightness synchrotron sources

  17. Analysis of the direct x-ray absorption noise in phosphor-coupled CMOS detectors

    International Nuclear Information System (INIS)

    Han, Jong Chul; Yun, Seung Man; Kim, Ho Kyung; Cunningham, Ian; Achterkirchen, Thorsten

    2009-01-01

    It is known that the indirect conversion detectors have an NPS (noise power spectrum), which decreases with the spatial frequency, and the direct conversion detector have a nearly constant NPS with the spatial frequency (or white NPS). This explains that when a significant amount of x rays are not absorbed in the phosphor layer, then the additional absorption of x-rays in the semiconductor layers (or the photodiodes) with their white noise contributions degrades the total NPS performance. From the fact, we investigated how the direct x-ray affects CMOS detectors in terms of NPS and DQE (detective quantum efficiency)

  18. Study on data acquisition circuit used in SSPA linear array detector X-ray detection

    CERN Document Server

    Wei Biao; Che Zhen Ping

    2002-01-01

    After SSPA used as X-ray array detector is developed, the authors take a research on the data acquisition circuit applied to the detector. The experiment designed has verified the feasibility of application of this array detector and its data acquisition circuit to X-ray computed tomography (X-CT). The preliminary test results indicate that the method of the X-ray detection is feasible for industry X-CT nondestructive testing, which brings about advantage for detecting and measuring with high resolution, good efficiency and low cost

  19. In-situ X-ray diffraction system using sources and detectors at fixed angular positions

    Science.gov (United States)

    Gibson, David M [Voorheesville, NY; Gibson, Walter M [Voorheesville, NY; Huang, Huapeng [Latham, NY

    2007-06-26

    An x-ray diffraction technique for measuring a known characteristic of a sample of a material in an in-situ state. The technique includes using an x-ray source for emitting substantially divergent x-ray radiation--with a collimating optic disposed with respect to the fixed source for producing a substantially parallel beam of x-ray radiation by receiving and redirecting the divergent paths of the divergent x-ray radiation. A first x-ray detector collects radiation diffracted from the sample; wherein the source and detector are fixed, during operation thereof, in position relative to each other and in at least one dimension relative to the sample according to a-priori knowledge about the known characteristic of the sample. A second x-ray detector may be fixed relative to the first x-ray detector according to the a-priori knowledge about the known characteristic of the sample, especially in a phase monitoring embodiment of the present invention.

  20. Development of an X-ray detector using surface plasmon resonance

    International Nuclear Information System (INIS)

    Kunieda, Y.; Nagashima, K.; Hasegawa, N.; Ochi, Y.

    2009-01-01

    A new X-ray detector using surface plasmon resonance (SPR) is proposed. The detector consists of a prism coated with a thin metal film and semiconductor film. Optical laser pulse induces SPR condition on the metal surface, and synchronized X-ray pulse which is absorbed into the semiconductor film can be detected by measuring the change of the resonance condition of the surface plasmon. The expected time and spatial resolution of this detector is better than that of conventional X-ray detectors by combining this SPR measurement with ultra-short laser pulse as the probe beam. Our preliminary investigation using Au and ZnSe coated prism implies this scheme works well as the detector for the ultra-short X-ray pulse.

  1. (Li) detector characteristics on the accuracy in X-ray analysis using the

    African Journals Online (AJOL)

    A study has been carried out to show how variations in Si(Li) detector characteristics affect the accuracy of X-ray spectra evaluation. The detector characteristics investigated are Be window thickness, Au layer, Si dead layer and Si Detector Sensitive volume. For each of the detector parameters, different thickness values ...

  2. Simulation of medical irradiation and X-ray detector signals

    Energy Technology Data Exchange (ETDEWEB)

    Kreisler, Bjoern

    2010-02-08

    -Carlo simulation of the irradiation head of the medical linear accelerator is constructed for the irradiation with electrons and photons in the second part of this thesis. The simulation is validated by comparison with water phantom measurements. Available information from the measurements is limited to the depth dose deposition and dose profiles in selected water depths. The simulation allows the evaluation of dose depositions and spectral particle distributions at various locations. In the electron irradiation mode, a collimation close to the patient is performed by an electron applicator. A possible reduction of the side leakage is evaluated by optimising the collimation of the beam. Medical irradiation with MeV-electrons and X-ray photons is the part of the clinical tumour treatment which is evaluated in this work. (orig.)

  3. Simulation of medical irradiation and X-ray detector signals

    International Nuclear Information System (INIS)

    Kreisler, Bjoern

    2010-01-01

    the irradiation head of the medical linear accelerator is constructed for the irradiation with electrons and photons in the second part of this thesis. The simulation is validated by comparison with water phantom measurements. Available information from the measurements is limited to the depth dose deposition and dose profiles in selected water depths. The simulation allows the evaluation of dose depositions and spectral particle distributions at various locations. In the electron irradiation mode, a collimation close to the patient is performed by an electron applicator. A possible reduction of the side leakage is evaluated by optimising the collimation of the beam. Medical irradiation with MeV-electrons and X-ray photons is the part of the clinical tumour treatment which is evaluated in this work. (orig.)

  4. High temperature GaAs X-ray detectors

    Science.gov (United States)

    Lioliou, G.; Whitaker, M. D. C.; Barnett, A. M.

    2017-12-01

    Two GaAs p+-i-n+ mesa X-ray photodiodes were characterized for their electrical and photon counting X-ray spectroscopic performance over the temperature range of 100 °C to -20 °C. The devices had 10 μm thick i layers with different diameters: 200 μm (D1) and 400 μm (D2). The electrical characterization included dark current and capacitance measurements at internal electric field strengths of up to 50 kV/cm. The determined properties of the two devices were compared with previously reported results that were made with a view to informing the future development of photon counting X-ray spectrometers for harsh environments, e.g., X-ray fluorescence spectroscopy of planetary surfaces in high temperature environments. The best energy resolution obtained (Full Width at Half Maximum at 5.9 keV) decreased from 2.00 keV at 100 °C to 0.66 keV at -20 °C for the spectrometer with D1, and from 2.71 keV at 100 °C to 0.71 keV at -20 °C for the spectrometer with D2. Dielectric noise was found to be the dominant source of noise in the spectra, apart from at high temperatures and long shaping times, where the main source of photopeak broadening was found to be the white parallel noise.

  5. Small angle X-ray scattering experiments with three-dimensional imaging gas detectors

    International Nuclear Information System (INIS)

    La Monaca, A.; Iannuzzi, M.; Messi, R.

    1985-01-01

    Measurements of small angle X-ray scattering of lupolen - R, dry collagen and dry cornea are presented. The experiments have been performed with synchrotron radiation and a new three-dimensional imaging drif-chamber gas detector

  6. Photovoltaic X-ray detectors based on epitaxial GaAs structures

    Energy Technology Data Exchange (ETDEWEB)

    Achmadullin, R.A. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Artemov, V.V. [Shubnikov Institute of Crystallography, Russian Academy of Sciences, 59 Leninski pr., Moscow B-333, 117333 (Russian Federation); Dvoryankin, V.F. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation)]. E-mail: vfd217@ire216.msk.su; Dvoryankina, G.G. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Dikaev, Yu.M. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Ermakov, M.G. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Ermakova, O.N. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Chmil, V.B. [Scientific State Center, High Energy Physics Institute, Protvino, Moscow region (Russian Federation); Holodenko, A.G. [Scientific State Center, High Energy Physics Institute, Protvino, Moscow region (Russian Federation); Kudryashov, A.A.; Krikunov, A.I.; Petrov, A.G.; Telegin, A.A. [Institute of Radio Engineering and Electronics, Russian Academy of Sciences, 1 Ac. Vvedenski square, Fryazino 141190, Moscow region (Russian Federation); Vorobiev, A.P. [Scientific State Center, High Energy Physics Institute, Protvino, Moscow region (Russian Federation)

    2005-12-01

    A new type of the photovoltaic X-ray detector based on epitaxial p{sup +}-n-n'-n{sup +} GaAs structures which provides a high efficiency of charge collection in the non-bias operation mode at room temperature is proposed. The GaAs epitaxial structures were grown by vapor-phase epitaxy on heavily doped n{sup +}-GaAs(1 0 0) substrates. The absorption efficiency of GaAs X-ray detector is discussed. I-V and C-V characteristics of the photovoltaic X-ray detectors are analyzed. The built-in electric field profiles in the depletion region of epitaxial structures are measured by the EBIC method. Charge collection efficiency to {alpha}-particles and {gamma}-radiation are measured. The application of X-ray detectors is discussed.

  7. Evaluation of In-Vacuum Imaging Plate Detector for X-Ray Diffraction Microscopy

    International Nuclear Information System (INIS)

    Nishino, Yoshinori; Takahashi, Yukio; Yamamoto, Masaki; Ishikawa, Tetsuya

    2007-01-01

    We performed evaluation tests of a newly developed in-vacuum imaging plate (IP) detector for x-ray diffraction microscopy. IP detectors have advantages over direct x-ray detection charge-coupled device (CCD) detectors, which have been commonly used in x-ray diffraction microscopy experiments, in the capabilities for a high photon count and for a wide area. The detector system contains two IPs to make measurement efficient by recording data with the one while reading or erasing the other. We compared speckled diffraction patterns of single particles taken with the IP and a direct x-ray detection CCD. The IP was inferior to the CCD in spatial resolution and in signal-to-noise ratio at a low photon count

  8. High gain gas microstrip detectors for soft x-ray detection

    International Nuclear Information System (INIS)

    Bateman, J.; Barlow, R.; Derbyshire, G.

    2001-01-01

    This report describes development work in which systematic changes in the pitch of the electrode pattern of a Gas Microstrip Detector are explored in the search for higher avalanche gains and enhanced stability. With the cathode width set to half of the pitch, gas gains of >50 000 are comfortably attainable with low detector noise so that x-rays can potentially be detected down to the limit of a single x-ray-produced photoelectron. (author)

  9. Comparison of experimental and theoretical efficiency of HPGe X-ray detector

    International Nuclear Information System (INIS)

    Mohanty, B.P.; Balouria, P.; Garg, M.L.; Nandi, T.K.; Mittal, V.K.; Govil, I.M.

    2008-01-01

    The low energy high purity germanium (HPGe) detectors are being increasingly used for the quantitative estimation of elements using X-ray spectrometric techniques. The softwares used for quantitative estimation normally evaluate model based efficiency of detector using manufacturer supplied detector physical parameters. The present work shows that the manufacturer supplied detector parameters for low energy HPGe detectors need to be verified by comparing model based efficiency with the experimental ones. This is particularly crucial for detectors with ion implanted P type contacts

  10. X-ray television area detectors for macromolecular structural studies with synchrotron radiation sources

    International Nuclear Information System (INIS)

    Arndt, U.W.; Gilmore, D.J.

    1978-01-01

    Two-dimensional X-ray diffraction patterns may be recorded quantitatively by means of X-ray-to-electron converters which are scanned in a television-type raster scan. Detectors of this type are capable of operating over the whole range of counting rates from very low to higher than those with which other types of converters can deal. The component parts of an X-ray television detector are examined and the limits to the precision of the measurements are analysed. (Auth.)

  11. Background in xenon filled X-ray detectors

    International Nuclear Information System (INIS)

    Feroci, M.; Costa, E.; Dwyer, J.; Ford, E.; Kaaret, P.; Rapisarda, M.; Soffitta, P.

    1995-01-01

    Xenon based gas mixtures have been often used in proportional counters for X-ray astronomy in order to achieve a good efficiency in the medium/high X-ray energy range. Proportional counters flown on past missions (i.e. HEAO1 and EXOSAT) filled with Xe-based mixtures have shown a higher residual background (after that all the rejection techniques have been applied) with respect to Ar-based ones, operating in the same energy band and in the same radiation environment. We show, by means of Monte Carlo simulations, analytical computations and laboratory measurements, that such difference can be mostly understood in terms of higher internal background production and lower pulse discrimination efficiency in Xe-based gas filling, with respect to Ar-based ones. (orig.)

  12. Pixel array detector for X-ray free electron laser experiments

    Energy Technology Data Exchange (ETDEWEB)

    Philipp, Hugh T., E-mail: htp2@cornell.edu [Department of Physics, Laboratory of Solid State Physics, Cornell University, Ithaca, NY 14853 (United States); Hromalik, Marianne [Electrical and Computer Engineering, SUNY Oswego, Oswego, NY 13126 (United States); Tate, Mark; Koerner, Lucas [Department of Physics, Laboratory of Solid State Physics, Cornell University, Ithaca, NY 14853 (United States); Gruner, Sol M. [Department of Physics, Laboratory of Solid State Physics, Cornell University, Ithaca, NY 14853 (United States); Wilson Laboratory, Cornell University, CHESS, Ithaca, NY 14853 (United States)

    2011-09-01

    X-ray free electron lasers (XFELs) promise to revolutionize X-ray science with extremely high peak brilliances and femtosecond X-ray pulses. This will require novel detectors to fully realize the potential of these new sources. There are many current detector development projects aimed at the many challenges of meeting the XFEL requirements . This paper describes a pixel array detector (PAD) that has been developed for the Coherent X-ray Imaging experiment at the Linac Coherent Light Source (LCLS) at the SLAC National Laboratory . The detector features 14-bit in-pixel digitization; a 2-level in-pixel gain setting that can be used to make an arbitrary 2-D gain pattern that is adaptable to a particular experiment; the ability to handle instantaneous X-ray flux rates of 10{sup 17} photons per second; and continuous frames rates in excess of 120 Hz. The detector uses direct detection of X-rays in a silicon diode. The charge produced by the diode is integrated in a pixilated application specific integrated circuit (ASIC) which digitizes collected holes with single X-ray photon capability. Each ASIC is 194x185 pixels, each pixel is 110{mu}mx110{mu}m on a side. Each pixel can detect up to 2500 X-rays per frame in low-gain mode, yet easily detects single photons at high-gain. Cooled, single-chip detectors have been built and meet all the required specifications. SLAC National Laboratory is engaged in constructing a tiled, multi-chip 1516x1516 pixel detector.

  13. Spectral correction algorithm for multispectral CdTe x-ray detectors

    DEFF Research Database (Denmark)

    Christensen, Erik D.; Kehres, Jan; Gu, Yun

    2017-01-01

    Compared to the dual energy scintillator detectors widely used today, pixelated multispectral X-ray detectors show the potential to improve material identification in various radiography and tomography applications used for industrial and security purposes. However, detector effects, such as charge...

  14. Application of MSM InP detectors to the measurement of pulsed X-ray radiation

    Czech Academy of Sciences Publication Activity Database

    Ryc, L.; Dobrzanski, L.; Dubecký, L.; Kaczmarczyk, J.; Pfeifer, Miroslav; Riesz, F.; Slysz, W.; Surma, B.

    2008-01-01

    Roč. 163, 4-6 (2008), 559-567 ISSN 1042-0150 R&D Projects: GA MŠk(CZ) LC528 Institutional research plan: CEZ:AV0Z10100523 Keywords : InP detector * X-ray detector * picosecond detector * laser plasma Subject RIV: BH - Optics, Masers, Lasers Impact factor: 0.415, year: 2008

  15. Energy dependence evaluation of a ZnO detector for diagnostic X-ray beam

    International Nuclear Information System (INIS)

    Valença, C.P.V.; Silveira, M.A.L.; Macedo, M.A.; Santos, M.A.

    2015-01-01

    In the last decades the international organizations of human health and radiation protection have recommended certain care for using X-ray as a diagnosis tool. The current concern is to avoid any type of radiological accident or overdose to the patient. This can be done assessing the parameters of the X-ray equipment and there are various types of detectors available for that: ionizing chamber, semiconductor devices, etc. These detectors must be calibrated so that they can be used for any energy range and such a procedure is correlated with what is called the energy dependence of the detector. In accordance with the stated requirements of IEC 61267, the standard radiation quality beams and irradiation conditions (RQRs) are the tools and techniques for calibrating diagnostic X-Ray instruments and detectors. The purpose of this work is to evaluate the behavior of the energy dependence of a detector fabricated from a zinc oxide (ZnO) nanofilm. A Pantak industrial X-ray equipment was used to generate the RQR radiation quality beams and test three ZnO detector samples. A 6430 sub-femto-ammeter, Keithley, was used to bias the ZnO detector and simultaneously perform the output readings. The results showed that the ZnO device has some increase in its sensitivity to the ionizing radiation as the X-ray effective energy decreases unlike other types of semiconductor electronic devices typically used as an X-ray detector. We can be concluded that, after calibration, the ZnO device can be used as a diagnostic X-ray detector. (author)

  16. Evaluation of the energy dependence of a zinc oxide nanofilm X-ray detector

    International Nuclear Information System (INIS)

    Valenca, C.P.V.; Silveira, M.A.L.; Macedo, M.A.; Santos, L.A.P

    2015-01-01

    International organizations of human health and radiation protection have recommended certain care for using of the X-ray as a diagnosis tool to avoid any type of radiological accident or overdose to the patient. This can be done assessing the parameters of the X-ray equipment and there are various types of detectors available for that: ionizing chamber, electronic semiconductor devices, etc. These detectors must be calibrated so that they can be used for any energy range and such a procedure is correlated with what is called the energy dependence of the detector. In accordance with the stated requirements of IEC 61267, the standard radiation quality beams and irradiation conditions (RQRs) are the tools and techniques for calibrating diagnostic X-Ray instruments and detectors. The purpose of this work is to evaluate the behavior of the energy dependence of a detector fabricated from a zinc oxide (ZnO) nanofilm. A Pantak industrial X-ray equipment was used to generate the RQR radiation quality beams and test three ZnO detector samples. A 6430 sub-femto-ammeter, Keithley, was used to bias the ZnO detector and simultaneously perform the output readings. The results showed that the ZnO device has some increase in its sensitivity to the ionizing radiation as the X-ray effective energy decreases unlike other types of semiconductor electronic devices typically used as an X-ray detector. We can conclude that the ZnO device can be used as a diagnostic X-ray detector with an appropriate calibration. (author)

  17. Microsecond-scale X-ray imaging with Controlled-Drift Detectors

    International Nuclear Information System (INIS)

    Castoldi, A.; Galimberti, A.; Guazzoni, C.; Rehak, P.; Strueder, L.

    2006-01-01

    The Controlled-Drift Detector is a fully-depleted silicon detector that allows 2-D position sensing and energy spectroscopy of X-rays in the range 0.5-20keV with excellent time resolution (few tens of μs) and limited readout channels. In this paper we review the Controlled-Drift Detector operating principle and we present the X-ray imaging and spectroscopic capabilities of Controlled Drift Detectors in microsecond-scale experiments and the more relevant applications fields

  18. Directly-deposited blocking filters for high-performance silicon x-ray detectors

    Science.gov (United States)

    Bautz, M.; Kissel, S.; Masterson, R.; Ryu, K.; Suntharalingam, V.

    2016-07-01

    Silicon X-ray detectors often require blocking filters to mitigate noise and out-of-band signal from UV and visible backgrounds. Such filters must be thin to minimize X-ray absorption, so direct deposition of filter material on the detector entrance surface is an attractive approach to fabrication of robust filters. On the other hand, the soft (E OD 6) care must be taken to prevent light from entering the sides and mounting surfaces of the detector. Our methods have been used to deposit filters on the detectors of the REXIS instrument scheduled to fly on OSIRIS-ReX later this year.

  19. Digital signal processors for cryogenic high-resolution x-ray detector readout

    International Nuclear Information System (INIS)

    Friedrich, Stephan; Drury, Owen B.; Bechstein, Sylke; Hennig, Wolfgang; Momayezi, Michael

    2003-01-01

    We are developing fast digital signal processors (DSPs) to read out superconducting high-resolution X-ray detectors with on-line pulse processing. For superconducting tunnel junction (STJ) detector read-out, the DSPs offer online filtering, rise time discrimination and pile-up rejection. Compared to analog pulse processing, DSP readout somewhat degrades the detector resolution, but improves the spectral purity of the detector response. We discuss DSP performance with our 9-channel STJ array for synchrotron-based high-resolution X-ray spectroscopy. (author)

  20. Versatile, reprogrammable area pixel array detector for time-resolved synchrotron x-ray applications

    Energy Technology Data Exchange (ETDEWEB)

    Gruner, Sol [Cornell Univ., Ithaca, NY (United States)

    2010-05-01

    The final technical report for DOE grant DE-SC0004079 is presented. The goal of the grant was to perform research, development and application of novel imaging x-ray detectors so as to effectively utilize the high intensity and brightness of the national synchrotron radiation facilities to enable previously unfeasible time-resolved x-ray research. The report summarizes the development of the resultant imaging x-ray detectors. Two types of detector platforms were developed: The first is a detector platform (called a Mixed-Mode Pixel Array Detector, or MM-PAD) that can image continuously at over a thousand images per second while maintaining high efficiency for wide dynamic range signals ranging from 1 to hundreds of millions of x-rays per pixel per image. Research on an even higher dynamic range variant is also described. The second detector platform (called the Keck Pixel Array Detector) is capable of acquiring a burst of x-ray images at a rate of millions of images per second.

  1. Position sensitive X-ray or X-ray detector and 3-D-tomography using same

    International Nuclear Information System (INIS)

    1975-01-01

    A fan-shaped beam of penetrating radiation, such as X-ray or γ-ray radiation, is directed through a slice of the body to be analyzed into a position sensitive detector for deriving a shadowgraph of transmission or absorption of the penetrating radiation by the body. A number of such shadowgraphs are obtained for different angles of rotation of the fan-shaped beam relative to the center of the slice being analyzed. The detected fan beam shadowgraph data is reordered into shadowgraph data corresponding to sets of parallel paths of radiation through the body. The reordered parallel path shadowgraph data is then convoluted in accordance with a 3-D reconstruction method by convolution in a computer to derive a 3-D reconstructed tomograph of the body under analysis. In a preferred embodiment, the position sensitive detector comprises a multiwire detector wherein the wires are arrayed parallel to the direction of the divergent penetrating rays to be detected. A focussed grid collimator is interposed between the body and the position sensitive detector for collimating the penetrating rays to be detected. The source of penetrating radiation is preferably a monochromatic source

  2. CVD diamond based soft X-ray detector with fast response

    International Nuclear Information System (INIS)

    Li Fang; Hou Lifei; Su Chunxiao; Yang Guohong; Liu Shenye

    2010-01-01

    A soft X-ray detector has been made with high quality chemical vapor deposited (CVD) diamond and the electrical structure of micro-strip. Through the measurement of response time on a laser with the pulse width of 10 ps, the full width at half maximum of the data got in the oscilloscope was 115 ps. The rise time of the CVD diamond detector was calculated to be 49 ps. In the experiment on the laser prototype facility, the signal got by the CVD diamond detector was compared with that got by a soft X-ray spectrometer. Both signals coincided well. The detector is proved to be a kind of reliable soft X-ray detector with fast response and high signal-to-noise ratio. (authors)

  3. Applications of 'edge-on' illuminated porous plate detectors for diagnostic X-ray imaging

    CERN Document Server

    Shikhaliev, P M

    2002-01-01

    Scanning X-ray imaging systems for non-invasive diagnostics have several advantages over conventional imaging systems using area detectors. They significantly reduce the detected scatter radiation, cover large areas and potentially provide high spatial resolution. Applications of one-dimensional gaseous detectors and 'edge-on' illuminated silicon strip detectors for scanning imaging systems are currently under intensive investigation. The purpose of this work is to investigate 'edge-on' illuminated Porous Plate (PP) detectors for applications in diagnostic X-ray imaging. MicroChannel Plate (MCP), which is a common type of PP, has previously been investigated as a detector in surface-on illumination mode for medical X-ray imaging. However, its detection efficiency was too low for medical imaging applications. In the present study, the PP are used in the 'edge-on' illumination mode. Furthermore, the structural parameters of different PP types are optimized to improve the detection efficiency in the diagnostic X...

  4. Application of energy dispersive X-ray spectrometers with semiconductor detectors in radiometric analyses

    International Nuclear Information System (INIS)

    Jugelt, P.; Schieckel, M.

    1983-01-01

    Problems and possibilities of applying semiconductor detector spectrometers in radiometric analyses are described. A summary of the state of the art and tendencies of device engineering and spectra evaluation is given. Liquid-nitrogen cooled Li-drifted Si-detectors and high-purity Ge-detectors are compared. Semiconductor detectors working at room temperature are under development. In this connection CdTe and HgI 2 semiconductor detectors are compared. The use of small efficient computers in the spectrometer systems stimulates the development of algorithms for spectra analyses and for determining the concentration. Fields of application of energy dispersive X-ray spectrometers are X-ray diffraction and X-ray macroanalysis in investigating the structure of extensive surface regions

  5. Design considerations for soft X-ray television imaging detectors

    International Nuclear Information System (INIS)

    Kalata, K.; Golub, L.

    1988-01-01

    Television sensors for X-rays can be coupled to converters and image intensifiers to obtain active areas, high flux capabilities, quantum efficiency, high time resolution, or ease of construction and operation that may not be obtained with a directly illuminated sensor. A general purpose system which makes use of these capabilities for a number of applications is decribed. Some of the performance characteristics of this type of system are examined, and the expected future developments for such systems are briefly addressed. 19 refs

  6. Note: Application of a pixel-array area detector to simultaneous single crystal x-ray diffraction and x-ray absorption spectroscopy measurements

    International Nuclear Information System (INIS)

    Sun, Cheng-Jun; Brewe, Dale L.; Heald, Steve M.; Zhang, Bangmin; Chen, Jing-Sheng; Chow, G. M.; Venkatesan, T.

    2014-01-01

    X-ray diffraction (XRD) and X-ray absorption spectroscopy (XAS) are two main x-ray techniques in synchrotron radiation facilities. In this Note, we present an experimental setup capable of performing simultaneous XRD and XAS measurements by the application of a pixel-array area detector. For XRD, the momentum transfer in specular diffraction was measured by scanning the X-ray energy with fixed incoming and outgoing x-ray angles. By selecting a small fixed region of the detector to collect the XRD signal, the rest of the area was available for collecting the x-ray fluorescence for XAS measurements. The simultaneous measurement of XRD and X-ray absorption near edge structure for Pr 0.67 Sr 0.33 MnO 3 film was demonstrated as a proof of principle for future time-resolved pump-probe measurements. A static sample makes it easy to maintain an accurate overlap of the X-ray spot and laser pump beam

  7. A new miniature microchannel plate X-ray detector for synchrotron radiation

    International Nuclear Information System (INIS)

    Rosemeier, R.G.; Green, R.E. Jr.

    1982-01-01

    A state-of-the-art microchannel plate detector has been developed which allows real time X-ray imaging of X-ray diffraction as well as radiographic phenomenon. Advantages of the device include a 50 mm X-ray input, length less than 4'', and a weight of less than 1 lb. Since the use of synchrotron radiation is greatly facilitated by the capability of remote viewing of X-ray diffraction or radiographic images in real time, a prototype electro-optical system has been designed which couples the X-ray microchannel plate detector with a solid state television camera. Advantages of the miniature, lightweight, X-ray synchrotron camera include a large 50 mm X-ray input window, an output signal that is available in both analog format for display on a television monitor and in digital format for computer processing, and a completely modular design which allows all the components to be exchanged for other components optimally suited for the desired applications. (orig.)

  8. Spectrometry of X-ray beams using Cadmium and Zinc Teluride detector

    International Nuclear Information System (INIS)

    Becker, Paulo Henriques Bastos

    1997-06-01

    Determination of X-ray spectra to be utilized for medical diagnostics is a complementary process to the development of procedures to be applied to the quality control of radiodiagnostics X-ray equipment. Until some years ago, that was only possible using Germanium or Silicon detectors. Both have an excellent resolution in this energy range, but present also some restrictions as there are high costs and the necessity of operating them at temperature of liquid Nitrogen, which is not always available at the measurement's place. Room temperature detectors like Cadmium Telluride and Mercury Iodine don't have these restrictions. They, however, have a lower resolution and incomplete collection of the charges produced by their interaction with radiation. With technological advance of crystal growth in general and new techniques like cooling the crystal with a Peltier cell and rise time discrimination circuits, today Cadmium Telluride detectors show a resolution very close to that from Germanium detectors. This work relates to the routine use of Cadmium and Zinc Telluride detectors for measuring X-ray spectra in loco of diagnostic X-ray units. It characterizes the properties of a commercially available detector and offers a model for stripping the measured pulse height distribution. It was also developed a collimator to allow the direct measurement of the beam. The model developed and the constructed set-up were applied to two X-ray tubes and the achieved spectra compared with some spectra available from the literature. (author)

  9. Diagnostic x-ray spectra measurements using a silicon surface barrier detector

    International Nuclear Information System (INIS)

    Pani, R.; Laitano, R.F.

    1987-01-01

    A silicon surface barrier detector having a low efficiency for x-ray is used to analyse diagnostic x-ray spectra. This characteristic is advantageous in overcoming experimental problems caused by high fluence rates typical of diagnostic x-ray beams. The pulse height distribution obtained with silicon surface barrier detectors is very different from the true photon spectra because of the presence of escaped Compton photons and the fact that detection efficiency falls abruptly when photon energy increases. A detailed analysis of the spurious effects involved in detection is made by a Monte Carlo method. A stripping procedure is described for implementation on a personal computer. The validity of this method is tested by comparison with experimental results obtained with a Ge detector. The spectra obtained with the Si detector are in fairly good agreement with the analogous spectra measured with a Ge detector. The advantages of using Si as opposed to Ge detectors in x-ray spectrometry are: its simplicity of use, its greater economy for use in routine diagnostic x-ray spectroscopy and the possibility that the stripping procedure can be implemented on a personal computer. (author)

  10. Use of silicon microstrip detectors in medical diagnostic x-rays

    International Nuclear Information System (INIS)

    Cabal Rodriguez, Ana Ester

    2004-11-01

    This work presents the development and characterization of a single photon counting system based on silicon microstrip detectors, used in High Energy Physics experiments, and on low noise multichannel readout electronics. The thesis evaluates the feasibility of dual energy X-ray imaging with silicon microstrip detectors to be applied on medical diagnosis. Dual energy mammographic and angiographic experimental tests have been performed using the developed counting systems proto types, properly phantoms and quasi-monochromatic X ray beams, obtained on a compact dichromatic source based on a conventional X-ray tube and a mosaic crystal. A Monte Carlo simulation of the performance of the experimental setup for dual X-ray imaging has also been carried out using MCNP-4C transport code. We obtained good agreement between MCNP results and the experimental data. (Author)

  11. Schottky x-ray detectors based on a bulk β-Ga2O3 substrate

    Science.gov (United States)

    Lu, Xing; Zhou, Leidang; Chen, Liang; Ouyang, Xiaoping; Liu, Bo; Xu, Jun; Tang, Huili

    2018-03-01

    β-Ga2O3 Schottky barrier diodes (SBDs) have been fabricated on a bulk (100) β-Ga2O3 substrate and tested as X-ray detectors in this study. The devices exhibited good rectification properties, such as a high rectification ratio and a close-to-unity ideality factor. A high photo-to-dark current ratio exceeding 800 was achieved for X-ray detection, which was mainly attributed to the low reverse leakage current in the β-Ga2O3 SBDs. Furthermore, transient response of the β-Ga2O3 X-ray detectors was investigated, and two different detection mechanisms, photovoltaic and photoconductive, were identified. The results imply the great potential of β-Ga2O3 based devices for X-ray detection.

  12. Comparison of natural and synthetic diamond X-ray detectors.

    Science.gov (United States)

    Lansley, S P; Betzel, G T; Metcalfe, P; Reinisch, L; Meyer, J

    2010-12-01

    Diamond detectors are particularly well suited for dosimetry applications in radiotherapy for reasons including near-tissue equivalence and high-spatial resolution resulting from small sensitive volumes. However, these detectors have not become commonplace due to high cost and poor availability arising from the need for high-quality diamond. We have fabricated relatively cheap detectors from commercially-available synthetic diamond fabricated using chemical vapour deposition. Here, we present a comparison of one of these detectors with the only commercially-available diamond-based detector (which uses a natural diamond crystal). Parameters such as the energy dependence and linearity of charge with dose were investigated at orthovoltage energies (50-250 kV), and dose-rate dependence of charge at linear accelerator energy (6 MV). The energy dependence of a synthetic diamond detector was similar to that of the natural diamond detector, albeit with slightly less variation across the energy range. Both detectors displayed a linear response with dose (at 100 kV) over the limited dose range used. The sensitivity of the synthetic diamond detector was 302 nC/Gy, compared to 294 nC/Gy measured for the natural diamond detector; however, this was obtained with a bias of 246.50 V compared to a bias of 61.75 V used for the natural diamond detector. The natural diamond detector exhibited a greater dependency on dose-rate than the synthetic diamond detector. Overall, the synthetic diamond detector performed well in comparison to the natural diamond detector.

  13. Instrumentations in x-ray plasma polarization spectroscopy. Crystal spectrometer, polarimeter and detectors for astronomical observations

    Energy Technology Data Exchange (ETDEWEB)

    Baronova, Elena O.; Stepanenko, Mikhail M. [RRC Kurchatov Institute, Nuclear Fusion Institute, Moscow (Russian Federation); Jakubowski, Lech [Soltan Institute for Nuclear Studies, Swierk-Otwock (Poland); Tsunemi, Hiroshi [Osaka Univ., Graduate School of Science, Osaka (Japan)

    2002-08-01

    This report discusses the various problems which are encountered when a crystal spectrometer is used for the purpose of observing polarized x-ray lines. A polarimeter is proposed based on the novel idea of using two series of equivalent atomic planes in a single crystal. The present status of the astronomical x-ray detection techniques are described with emphasis on two dimensional detectors which are polarization sensitive. (author)

  14. Energy dependent features of X-ray signals in a GridPix detector

    Science.gov (United States)

    Krieger, C.; Kaminski, J.; Vafeiadis, T.; Desch, K.

    2018-06-01

    We report on the calibration of an argon/isobutane (97.7%/2.3%)-filled GridPix detector with soft X-rays (277 eV to 8 keV) using the variable energy X-ray source of the CAST Detector Lab at CERN. We study the linearity and energy resolution of the detector using both the number of pixels hit and the total measured charge as energy measures. For the latter, the energy resolution σE / E is better than 10% (20%) for energies above 2 keV (0.5 keV). Several characteristics of the recorded events are studied.

  15. Optimising the design of gas microstrip detectors for soft x-ray detection

    International Nuclear Information System (INIS)

    Bateman, J.; Barlow, R.; Derbyshire, G.

    2001-01-01

    This report describes development work in which systematic changes in the electrode pattern of a Gas Microstrip Detector are explored in the search for higher avalanche gains and enhanced stability. It is found that the width of the cathode structure is the main determinant of the detector stability. With the correct cathode width, gas gains of >50 000 are comfortably attainable with low detector noise so that x-rays can potentially be detected down to the limit of a single x-ray-produced photoelectron. (author)

  16. A high-resolution multiwire area detector for X-ray scattering

    Energy Technology Data Exchange (ETDEWEB)

    Faruqi, A R; Andrews, H [Medical Research Council, Cambridge (UK). Lab. of Molecular Biology

    1989-11-10

    A high-resolution multiwire area detector has been developed for recording X-ray scattering from biological specimens. The detector is 100x100 mm{sup 2} and, under the present operating conditions, has a spatial resolution of about 250 {mu}m in both directions. The detector is set up on a double-mirror focusing camera on a rotating anode X-ray generator and has been used in a number of small-angle experiments, two of which are described in this paper. (orig.).

  17. Adjustment of a low energy, X-rays generator (6 kV - 50 mA). Application to X-rays detectors calibration

    International Nuclear Information System (INIS)

    Legistre, C.

    1995-02-01

    The aim of this memoir is the calibration of an aluminium photocathode X-rays photoelectric detector, in the spectral range 0,5 keV - 1,5 KeV, with a continuous X-ray source. The detectors's calibration consist to measure the detector's sensitivity versus incident energy. In order to produce monochromatic incident beam on the detector, we used a multilayer mirror whose reflectivity was characterized. The measurements are compared to those realized in an other laboratory. (authors). 36 refs., 61 figs., 13 tabs., 2 photos

  18. Development of a lens-coupled CMOS detector for an X-ray inspection system

    International Nuclear Information System (INIS)

    Kim, Ho Kyung; Ahn, Jung Keun; Cho, Gyuseong

    2005-01-01

    A digital X-ray imaging detector based on a complementary metal-oxide-semiconductor (CMOS) image sensor has been developed for X-ray non-destructive inspection applications. This is a cost-effective solution because of the availability of cheap commercial standard CMOS image sensors. The detector configuration adopts an indirect X-ray detection method by using scintillation material and lens assembly. As a feasibility test of the developed lens-coupled CMOS detector as an X-ray inspection system, we have acquired X-ray projection images under a variety of imaging conditions. The results show that the projected image is reasonably acceptable in typical non-destructive testing (NDT). However, the developed detector may not be appropriate for laminography due to a low light-collection efficiency of lens assembly. In this paper, construction of the lens-coupled CMOS detector and its specifications are described, and the experimental results are presented. Using the analysis of quantum accounting diagram, inefficiency of the lens-coupling method is discussed

  19. Calibration of the hard x-ray detectors for the FOXSI solar sounding rocket

    Science.gov (United States)

    Athiray, P. S.; Buitrago-Casas, Juan Camilo; Bergstedt, Kendra; Vievering, Juliana; Musset, Sophie; Ishikawa, Shin-nosuke; Glesener, Lindsay; Takahashi, Tadayuki; Watanabe, Shin; Courtade, Sasha; Christe, Steven; Krucker, Säm.; Goetz, Keith; Monson, Steven

    2017-08-01

    The Focusing Optics X-ray Solar Imager (FOXSI) sounding rocket experiment conducts direct imaging and spectral observation of the Sun in hard X-rays, in the energy range 4 to 20 keV. These high-sensitivity observations are used to study particle acceleration and coronal heating. FOXSI is designed with seven grazing incidence optics modules that focus X-rays onto seven focal plane detectors kept at a 2m distance. FOXSI-1 was flown with seven Double-sided Si Strip Detectors (DSSD), and two of them were replaced with CdTe detectors for FOXSI-2. The upcoming FOXSI-3 flight will carry DSSD and CdTe detectors with upgraded optics for enhanced sensitivity. The detectors are calibrated using various radioactive sources. The detector's spectral response matrix was constructed with diagonal elements using a Gaussian approximation with a spread (sigma) that accounts for the energy resolution of the detector. Spectroscopic studies of past FOXSI flight data suggest that the inclusion of lower energy X-rays could better constrain the spectral modeling to yield a more precise temperature estimation of the hot plasma. This motivates us to carry out an improved calibration to better understand the finer-order effects on the spectral response, especially at lower energies. Here we report our improved calibration of FOXSI detectors using experiments and Monte-Carlo simulations.

  20. Si(Li) detector system for application to x-ray astronomy rocket experiments

    International Nuclear Information System (INIS)

    Griffiths, R.E.; Cheron, C.; Friant, A.; Jehanno, C.; Rocchia, R.; Rothenflug, R.; Testard, O.

    1975-01-01

    The problems associated with the use of Si(Li) detectors in x-ray astronomy rocket experiments are discussed. In particular a detector system is described that can be used at the focus of a grazing-incidence paraboloid telescope for the energy range 0.3 to 2 keV. (U.S.)

  1. Semiconductor detectors in current energy dispersive X-ray spectral analysis

    International Nuclear Information System (INIS)

    Betin, J.; Zhabin, E.; Krampit, I.; Smirnov, V.

    1980-01-01

    A review is presented of the properties of semiconductor detectors and of the possibilities stemming therefrom of using the detectors in X-ray spectral analysis in industries, in logging, in ecology and environmental control, in medicine, etc. (M.S.)

  2. Some aspects of detectors and electronics for x-ray fluorescence analysis

    International Nuclear Information System (INIS)

    Goulding, F.S.

    1976-08-01

    Some of the less recognized and potentially important parameters of the electronics and detectors used in X-ray fluorescence spectrometers are discussed. Detector factors include window (dead-layer) effects, time-dependent background and excess background. Noise parameters of field-effect transistors and time-variant pulse shaping are also discussed

  3. Linearity discontinuities in Xe-filled X-ray microstrip detectors

    DEFF Research Database (Denmark)

    Zavattini, G.; Feroci, M.; Budtz-Jørgensen, Carl

    1997-01-01

    A prototype Xe + 10% CH4 microstrip detector was used to study the K-edge discontinuity in the pulse-height distribution as a function of the energy of incident X-rays. The electronics used was such that a pulse-shape rejection could be made of K-fluorescence reabsorption in the detector. The mea...

  4. X-ray detectors for structure investigations constructed at JINR

    Energy Technology Data Exchange (ETDEWEB)

    Chernenko, S P [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Cheremukhina, G A [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Fateev, O V [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Smykov, L P [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Vasiliev, S E [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Zanevsky, Yu V [LHE Joint Institute for Nuclear Research, 141980 Dubna (Russian Federation); Kheiker, D M [Institute of Crystallography, Leninsky prosp. 59, 117333 Moscow (Russian Federation); Popov, A N [Institute of Crystallography, Leninsky prosp. 59, 117333 Moscow (Russian Federation)

    1994-09-01

    The performance characteristics of a few high resolution position-sensitive detectors constructed at JINR are presented. The detectors supplied with original software operate with an IBM PC/AT. One of these devices has been succesfully applied for protein molecule structure investigations and the other for studies of the structure-forming process during combustion. The preliminary parameters of the high count rate MWPC with parallel electronics and the testing results of the microstrip detector are given. ((orig.))

  5. A semiempirical linear model of indirect, flat-panel x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M. [Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Psychological and Brain Sciences, University of California, Santa Barbara, California 92106 (United States); Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 3100, Sacramento, California 95817 (United States)

    2012-04-15

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r

  6. A semiempirical linear model of indirect, flat-panel x-ray detectors

    International Nuclear Information System (INIS)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M.

    2012-01-01

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r 2 of

  7. A semiempirical linear model of indirect, flat-panel x-ray detectors.

    Science.gov (United States)

    Huang, Shih-Ying; Yang, Kai; Abbey, Craig K; Boone, John M

    2012-04-01

    It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV∕0.1-mm Sn, 65 kV∕0.2-mm Cu, 85 kV∕1.5-mm Al, and 95 kV∕0.05-mm Ag. The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r(2) of signal mean to k

  8. Recent developments in detectors/phantoms for dosimetry, X-ray quality assurance and imaging

    International Nuclear Information System (INIS)

    Sankaran, A.

    2009-01-01

    During the past years, many new developments have taken place in detectors/phantoms for high energy photon and electron dosimetry (for radiotherapy), protection monitoring, X-ray quality assurance and X-ray imaging (for radiodiagnosis). A variety of detectors and systems, quality assurance (QA) gadgets and special phantoms have been developed for diverse applications. This paper discusses the important developments with some of which the author was actively associated in the past. For dosimetry and QA of 60 Co and high energy X-ray units, state-of-the-art radiation field analyzers, matrix ion chambers, MOSFET devices and Gafchromic films are described. OSL detectors find wide use in radiotherapy dosimetry and provide a good alternative for personnel monitoring. New systems introduced for QA/dosimetry of X-ray units and CT scanners include: multi-function instruments for simultaneous measurement of kVp, dose, time, X-ray waveform and HVT on diagnostic X-ray units; pencil chamber with head and body phantoms for CTDI check on CT scanners. Examples of phantoms used for dosimetry and imaging are given. Advancements in the field of diagnostic X-ray imaging (with applications in portal imaging/dosimetry of megavoltage X-ray units) have led to emergence of: film-replacement systems employing CCD-scintillator arrays, computed radiography (CR) using storage phosphor plate; digital radiography (DR), using a pixel-matrix of amorphous selenium, or amorphous silicon diode coupled to scintillator. All these provide (a) in radiotherapy, accurate dose delivery to tumour, saving the surrounding tissues and (b) in radiodiagnosis, superior image quality with low patient exposure. Lastly, iPODs and flash drives are utilized for storage of gigabyte-size images encountered in medical and allied fields. Although oriented towards medical applications, some of these have been of great utility in other fields, such as industrial radiography as well as a host of other research areas. (author)

  9. New portable 1200 W X-ray detectors

    International Nuclear Information System (INIS)

    Liebram, A.; Niemann, W.; Nielsen, J.

    2005-01-01

    The contribution presents new detectors of up to 1200 W. The new detectors by YXLON have several of the following advantages, depending on the application: a) Reduced exposure time; b) Higher contrast; c) deeper penetration. The high efficiency, the duty cycle achieved, and the low price make the instruments interesting also for stationary applications in radiography and radioscopy [de

  10. Correction method and software for image distortion and nonuniform response in charge-coupled device-based x-ray detectors utilizing x-ray image intensifier

    International Nuclear Information System (INIS)

    Ito, Kazuki; Kamikubo, Hironari; Yagi, Naoto; Amemiya, Yoshiyuki

    2005-01-01

    An on-site method of correcting the image distortion and nonuniform response of a charge-coupled device (CCD)-based X-ray detector was developed using the response of the imaging plate as a reference. The CCD-based X-ray detector consists of a beryllium-windowed X-ray image intensifier (Be-XRII) and a CCD as the image sensor. An image distortion of 29% was improved to less than 1% after the correction. In the correction of nonuniform response due to image distortion, subpixel approximation was performed for the redistribution of pixel values. The optimal number of subpixels was also discussed. In an experiment with polystyrene (PS) latex, it was verified that the correction of both image distortion and nonuniform response worked properly. The correction for the 'contrast reduction' problem was also demonstrated for an isotropic X-ray scattering pattern from the PS latex. (author)

  11. PILATUS: a two-dimensional X-ray detector for macromolecular crystallography

    CERN Document Server

    Eikenberry, E F; Huelsen, G; Toyokawa, H; Horisberger, R P; Schmitt, B; Schulze-Briese, C; Tomizaki, T

    2003-01-01

    A large quantum-limited area X-ray detector for protein crystallography is under development at the Swiss Light Source. The final detector will be 2kx2k pixels covering 40x40 cm sup 2. A three-module prototype with 1120x157 pixels covering an active area of 24.3x3.4 cm sup 2 has been tested. X-rays above 6 keV with peak count rates exceeding 5x10 sup 5 X-ray/pixel/s could be detected in single photon counting mode. Statistics of module production and results of threshold trimming are presented. To demonstrate the potential of this new detector, protein crystal data were collected at beamline 6S of the SLS.

  12. X-ray imaging with photon counting hybrid semiconductor pixel detectors

    CERN Document Server

    Manolopoulos, S; Campbell, M; Snoeys, W; Heijne, Erik H M; Pernigotti, E; Raine, C; Smith, K; Watt, J; O'Shea, V; Ludwig, J; Schwarz, C

    1999-01-01

    Semiconductor pixel detectors, originally developed for particle physics experiments, have been studied as X-ray imaging devices. The performance of devices using the OMEGA 3 read-out chip bump-bonded to pixellated silicon semiconductor detectors is characterised in terms of their signal-to-noise ratio when exposed to 60 kVp X-rays. Although parts of the devices achieve values of this ratio compatible with the noise being photon statistics limited, this is not found to hold for the whole pixel matrix, resulting in the global signal-to-noise ratio being compromised. First results are presented of X-ray images taken with a gallium arsenide pixel detector bump-bonded to a new read-out chip, (MEDIPIX), which is a single photon counting read-out chip incorporating a 15-bit counter in every pixel. (author)

  13. Two-dimensional imaging detectors for structural biology with X-ray lasers.

    Science.gov (United States)

    Denes, Peter

    2014-07-17

    Our ability to harness the advances in microelectronics over the past decade(s) for X-ray detection has resulted in significant improvements in the state of the art. Biology with X-ray free-electron lasers present daunting detector challenges: all of the photons arrive at the same time, and individual high peak power pulses must be read out shot-by-shot. Direct X-ray detection in silicon pixel detectors--monolithic or hybrid--are the standard for XFELs today. For structural biology, improvements are needed for today's 10-100 Hz XFELs, and further improvements are required for tomorrow's 10+ kHz XFELs. This article will discuss detector challenges, why they arise and ways to overcome them, along with the current state of the art. © 2014 The Author(s) Published by the Royal Society. All rights reserved.

  14. Characterization of LiF-based soft X-ray imaging detectors by confocal fluorescence microscopy

    International Nuclear Information System (INIS)

    Bonfigli, F; Gaudio, P; Lupelli, I; Nichelatti, E; Richetta, M; Vincenti, M A; Montereali, R M

    2010-01-01

    X-ray microscopy represents a powerful tool to obtain images of samples with very high spatial resolution. The main limitation of this technique is represented by the poor spatial resolution of standard imaging detectors. We proposed an innovative high-performance X-ray imaging detector based on the visible photoluminescence of colour centres in lithium fluoride. In this work, a confocal microscope in fluorescence mode was used to characterize LiF-based imaging detectors measuring CC integrated visible fluorescence signals of LiF crystals and films (grown on several kinds of substrates) irradiated by soft X-rays produced by a laser plasma source in different exposure conditions. The results are compared with the CC photoluminescence spectra measured on the same samples and discussed.

  15. Flexible X-ray detector based on sliced lead iodide crystal

    Energy Technology Data Exchange (ETDEWEB)

    Sun, Hui; Gao, Xiuying [College of Optoelectronic Technology, Chengdu University of Information Technology, Chengdu (China); Department of Materials Science, Sichuan University, Chengdu (China); Zhao, Beijun [Department of Materials Science, Sichuan University, Chengdu (China); Yang, Dingyu; Wangyang, Peihua; Zhu, Xinghua [College of Optoelectronic Technology, Chengdu University of Information Technology, Chengdu (China)

    2017-02-15

    A promising flexible X-ray detector based on inorganic semiconductor PbI{sub 2} crystal is reported. The sliced crystals mechanically cleaved from an as-grown PbI{sub 2} crystal act as the absorber directly converting the impinging X-ray photons to electron hole pairs. Due to the ductile feature of the PbI{sub 2} crystal, the detector can be operated under a highly curved state with the strain on the top surface up to 1.03% and still maintaining effective detection performance. The stable photocurrent and fast response were obtained with the detector repeated bending to a strain of 1.03% for 100 cycles. This work presents an approach for developing efficient and cost-effective PbI{sub 2}-based flexible X-ray detector. Photocurrent responses of the flexible PbI{sub 2} X-ray detector with the strain on the top surface up to 1.03% proposed in this work with the cross sectional structure and curved detector photograph as insets. (copyright 2016 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim)

  16. Photovoltaic x-ray detectors based on the GaAs epitaxial structures

    CERN Document Server

    Akhmadullin, R A; Dvoryankina, G G; Dikaev, Y M; Ermakov, M G; Ermakova, O N; Krikunov, A I; Kudryashov, A A; Petrov, A G; Telegin, A A

    2002-01-01

    The new photovoltaic detector of the X-ray radiation is proposed on the basis of the GaAs epitaxial structures, which operates with high efficiency of the charge carriers collection without shift voltage and at the room temperature. The structures are grown by the method of the gas-phase epitaxy on the n sup + -type highly-alloyed substrates. The range of sensitivity to the X-ray radiation is within the range of effective energies from 8 up to 120 keV. The detector maximum response in the current short circuit mode is determined

  17. Plastic nuclear track detectors as high x-ray and gamma dosimeters

    International Nuclear Information System (INIS)

    Chong Chon Sing

    1995-01-01

    A brief review of recent studies on the effects of high doses of x-ray and gamma ray on the track registration properties of several plastic track detectors is presented. The bulk etching rates and the etched track sizes have been found to increase with the dose in the range up to 100 Mrad. These results suggest that the changes in track registration characteristics can be employed as an index of the radiation dose in the megarad region. In particular, recent results on the effect of X-ray irradiation on two types of cellulose nitrate track detectors obtained in our laboratory are reported in this paper. (author)

  18. X-ray system with coupled source drive and detector drive

    International Nuclear Information System (INIS)

    1976-01-01

    An electronic coupling replacing the (more expensive) mechanical coupling which controls the speed of two sets of two electric motors, one driving an X-ray source and the other an X-ray detector, is described. Source and detector are kept rotating in parallel planes with a fairly constant velocity ratio. The drives are controlled by an electronic system comprising a comparator circuit comparing the position-indicative signals, a process control circuit and an inverter switch. The control system regulates the speed of the electric motors. The signal processing is described

  19. Development of CdZnTe X-ray detectors at DSRI

    DEFF Research Database (Denmark)

    van Pamelen, M.A.J.; Budtz-Jørgensen, Carl; Kuvvetli, Irfan

    2000-01-01

    An overview of the development of CdZnTe X-ray detectors at the Danish Space Research Institute is presented. Initiated in the beginning of 1996, the main motivation at that time was to develop focal plane detectors for the novel type of hard X-ray telescopes, which are currently under study...... developed a technique, which, with the use of microstrip electrodes, is able to compensate for the signal loss caused by trapping of positive charge carriers. This technique leads to a dramatic improvement of the achievable energy resolution, even for crystals of poor quality. With the technique, hole...

  20. Developments in microchannel plate detectors for imaging x-ray astronomy

    International Nuclear Information System (INIS)

    Fraser, G.W.; Whiteley, M.J.; Pearson, J.F.

    1985-01-01

    The authors present new results in four areas of microchannel plate (MCP) X-ray detector operation. The performance in pulse counting mode of MCPs with 8 micron channel diameters is reported. The effects on MCP quantum detection efficiency and energy discrimination of multiple CsI coatings are described. A new mode of operation of two-stage multipliers is evaluated. Replacing the conventional electron-accelerating inter-plate potential difference by a retarding field is shown to result in definite advantages with regard to X-ray energy discrimination and detector lifetime. The source of the MCP internal background is discussed

  1. Development of a fast multi-line x-ray CT detector for NDT

    International Nuclear Information System (INIS)

    Hofmann, T.; Nachtrab, F.; Schlechter, T.; Mühlbauer, J.; Schröpfer, S.; Firsching, M.; Uhlmann, N.; Neubauer, H.; Ernst, J.; Schweiger, T.; Oberst, M.; Meyer, A.

    2015-01-01

    Typical X-ray detectors for non-destructive testing (NDT) are line detectors or area detectors, like e.g. flat panel detectors. Multi-line detectors are currently only available in medical Computed Tomography (CT) scanners. Compared to flat panel detectors, line and multi-line detectors can achieve much higher frame rates. This allows time-resolved 3D CT scans of an object under investigation. Also, an improved image quality can be achieved due to reduced scattered radiation from object and detector themselves. Another benefit of line and multi-line detectors is that very wide detectors can be assembled easily, while flat panel detectors are usually limited to an imaging field with a size of approx. 40 × 40 cm 2 at maximum. The big disadvantage of line detectors is the limited number of object slices that can be scanned simultaneously. This leads to long scan times for large objects. Volume scans with a multi-line detector are much faster, but with almost similar image quality. Due to the promising properties of multi-line detectors their application outside of medical CT would also be very interesting for NDT. However, medical CT multi-line detectors are optimized for the scanning of human bodies. Many non-medical applications require higher spatial resolutions and/or higher X-ray energies. For those non-medical applications we are developing a fast multi-line X-ray detector.In the scope of this work, we present the current state of the development of the novel detector, which includes several outstanding properties like an adjustable curved design for variable focus-detector-distances, conserving nearly uniform perpendicular irradiation over the entire detector width. Basis of the detector is a specifically designed, radiation hard CMOS imaging sensor with a pixel pitch of 200 μ m. Each pixel has an automatic in-pixel gain adjustment, which allows for both: a very high sensitivity and a wide dynamic range. The final detector is planned to have 256 lines of

  2. Compton polarimetry with position-resolving X-ray detectors

    International Nuclear Information System (INIS)

    Hess, Sebastian

    2010-02-01

    In the present thesis the prototype of a novel position-resolving and multi-hit able 2D Si(Li) strip detector is characterized, the planar detector crystal of which is simultaneously applied both as scatterer and as absorber. In the framework of this thesis the Si(Li) polarimeter could be applied in different experiments on the radiative electron capture and on the characteristic radiation at the experimental storage ring of the GSI. The characterization of the detector pursued by means of the highly polarized radiation of the electron capture into the K shell of naked xenon. In the following in two further experiments new values on the polarization of the electron capture into the K shell both of the naked and of the hydrogen-like uranium were performed.

  3. Advances in indirect detector systems for ultra high-speed hard X-ray imaging with synchrotron light

    Science.gov (United States)

    Olbinado, M. P.; Grenzer, J.; Pradel, P.; De Resseguier, T.; Vagovic, P.; Zdora, M.-C.; Guzenko, V. A.; David, C.; Rack, A.

    2018-04-01

    We report on indirect X-ray detector systems for various full-field, ultra high-speed X-ray imaging methodologies, such as X-ray phase-contrast radiography, diffraction topography, grating interferometry and speckle-based imaging performed at the hard X-ray imaging beamline ID19 of the European Synchrotron—ESRF. Our work highlights the versatility of indirect X-ray detectors to multiple goals such as single synchrotron pulse isolation, multiple-frame recording up to millions frames per second, high efficiency, and high spatial resolution. Besides the technical advancements, potential applications are briefly introduced and discussed.

  4. Advancing the technology of monolithic CMOS detectors for use as x-ray imaging spectrometers

    Science.gov (United States)

    Kenter, Almus; Kraft, Ralph; Gauron, Thomas; Amato, Stephen

    2017-08-01

    The Smithsonian Astrophysical Observatory (SAO) in collaboration with SRI/Sarnoff has been engaged in a multi year effort to advance the technology of monolithic back-thinned CMOS detectors for use as X-ray imaging spectrometers. The long term goal of this campaign is to produce X-ray Active Pixel Sensor (APS) detectors with Fano limited performance over the 0.1-10keV band while incorporating the many benefits of CMOS technology. These benefits include: low power consumption, radiation "hardness", high levels of integration, and very high read rates. Such devices would be ideal for candidate post 2020 decadal missions such as LYNX and for smaller more immediate applications such as CubeX. Devices from a recent fabrication have been back-thinned, packaged and tested for soft X-ray response. These devices have 16μm pitch, 6 Transistor Pinned Photo Diode (6TPPD) pixels with ˜135μV/electron sensitivity and a highly parallel signal chain. These new detectors are fabricated on 10μm epitaxial silicon and have a 1k by 1k format. We present details of our camera design and device performance with particular emphasis on those aspects of interest to single photon counting X-ray astronomy. These features include read noise, X-ray spectral response and quantum efficiency.

  5. Bismuth Passivation Technique for High-Resolution X-Ray Detectors

    Science.gov (United States)

    Chervenak, James; Hess, Larry

    2013-01-01

    The Athena-plus team requires X-ray sensors with energy resolution of better than one part in 3,000 at 6 keV X-rays. While bismuth is an excellent material for high X-ray stopping power and low heat capacity (for large signal when an X-ray is stopped by the absorber), oxidation of the bismuth surface can lead to electron traps and other effects that degrade the energy resolution. Bismuth oxide reduction and nitride passivation techniques analogous to those used in indium passivation are being applied in a new technique. The technique will enable improved energy resolution and resistance to aging in bismuth-absorber-coupled X-ray sensors. Elemental bismuth is lithographically integrated into X-ray detector circuits. It encounters several steps where the Bi oxidizes. The technology discussed here will remove oxide from the surface of the Bi and replace it with nitridized surface. Removal of the native oxide and passivating to prevent the growth of the oxide will improve detector performance and insulate the detector against future degradation from oxide growth. Placing the Bi coated sensor in a vacuum system, a reduction chemistry in a plasma (nitrogen/hydrogen (N2/H2) + argon) is used to remove the oxide and promote nitridization of the cleaned Bi surface. Once passivated, the Bi will perform as a better X-ray thermalizer since energy will not be trapped in the bismuth oxides on the surface. A simple additional step, which can be added at various stages of the current fabrication process, can then be applied to encapsulate the Bi film. After plasma passivation, the Bi can be capped with a non-diffusive layer of metal or dielectric. A non-superconducting layer is required such as tungsten or tungsten nitride (WNx).

  6. Picosecond time-resolved laser pump/X-ray probe experiments using a gated single-photon-counting area detector

    DEFF Research Database (Denmark)

    Ejdrup, T.; Lemke, H.T.; Haldrup, Martin Kristoffer

    2009-01-01

    The recent developments in X-ray detectors have opened new possibilities in the area of time-resolved pump/probe X-ray experiments; this article presents the novel use of a PILATUS detector to achieve X-ray pulse duration limited time-resolution at the Advanced Photon Source (APS), USA...... limited time-resolution of 60 ps using the gated PILATUS detector. This is the first demonstration of X-ray pulse duration limited data recorded using an area detector without the use of a mechanical chopper array at the beamline........ The capability of the gated PILATUS detector to selectively detect the signal from a given X-ray pulse in 24 bunch mode at the APS storage ring is demonstrated. A test experiment performed on polycrystalline organic thin films of [alpha]-perylene illustrates the possibility of reaching an X-ray pulse duration...

  7. Extra-oral dental radiography for disaster victims using a flat panel X-ray detector and a hand-held X-ray generator.

    Science.gov (United States)

    Ohtani, M; Oshima, T; Mimasaka, S

    2017-12-01

    Forensic odontologists commonly incise the skin for post-mortem dental examinations when it is difficult to open the victim's mouth. However, it is prohibited by law to incise dead bodies without permission in Japan. Therefore, we attempted using extra-oral dental radiography, using a digital X-ray equipment with rechargeable batteries, to overcome this restriction. A phantom was placed in the prone position on a table, and three plain dental radiographs were used per case: "lateral oblique radiographs" for left and right posterior teeth and a "contact radiograph" for anterior teeth were taken using a flat panel X-ray detector and a hand-held X-ray generator. The resolving power of the images was measured by a resolution test chart, and the scattered X-ray dose was measured using an ionization chamber-type survey meter. The resolving power of the flat panel X-ray detector was 3.0 lp/mm, which was less than that of intra-oral dental methods, but the three extra-oral plain dental radiographs provided the overall dental information from outside of the mouth, and this approach was less time-consuming. In addition, the higher dose of scattered X-rays was laterally distributed, but the dose per case was much less than that of intra-oral dental radiographs. Extra-oral plain dental radiography can be used for disaster victim identification by dental methods even when it is difficult to open the mouth. Portable and rechargeable devices, such as a flat panel X-ray detector and a hand-held X-ray generator, are convenient to bring and use anywhere, even at a disaster scene lacking electricity and water.

  8. High-speed X-ray imaging pixel array detector for synchrotron bunch isolation.

    Science.gov (United States)

    Philipp, Hugh T; Tate, Mark W; Purohit, Prafull; Shanks, Katherine S; Weiss, Joel T; Gruner, Sol M

    2016-03-01

    A wide-dynamic-range imaging X-ray detector designed for recording successive frames at rates up to 10 MHz is described. X-ray imaging with frame rates of up to 6.5 MHz have been experimentally verified. The pixel design allows for up to 8-12 frames to be stored internally at high speed before readout, which occurs at a 1 kHz frame rate. An additional mode of operation allows the integration capacitors to be re-addressed repeatedly before readout which can enhance the signal-to-noise ratio of cyclical processes. This detector, along with modern storage ring sources which provide short (10-100 ps) and intense X-ray pulses at megahertz rates, opens new avenues for the study of rapid structural changes in materials. The detector consists of hybridized modules, each of which is comprised of a 500 µm-thick silicon X-ray sensor solder bump-bonded, pixel by pixel, to an application-specific integrated circuit. The format of each module is 128 × 128 pixels with a pixel pitch of 150 µm. In the prototype detector described here, the three-side buttable modules are tiled in a 3 × 2 array with a full format of 256 × 384 pixels. The characteristics, operation, testing and application of the detector are detailed.

  9. Upgrade of the InGrid based X-ray detector for the CAST experiment

    Energy Technology Data Exchange (ETDEWEB)

    Desch, Klaus; Kaminski, Jochen; Krieger, Christoph; Schmidt, Sebastian [Physikalisches Institut, Universitaet Bonn, Nussallee 12, 53115 Bonn (Germany)

    2016-07-01

    The CERN Axion Solar Telescope (CAST) is a magnetic helioscope searching for solar axions and chameleons using the inverse Primakoff effect. The produced photons are in the low X-ray regime. Chameleon search demands high sensitivity to photons with less than 1 keV and a very low background rate. Several improvements to the detector design used in 2014/15 are envisaged for 2016. The readout system is to be improved by including a flash ADC to read out the analog signal induced on the grid. The pulse shape contains information about the longitudinal shape of the event in addition to the transverse shape given by the pixel read out. Tracks passing through the chip orthogonally resemble photons in transverse shape. A scintillator behind the detector will also allow cross referencing chip and and scintillator signals to further reduce background rates. Finally, a new X-ray window separating detector and X-ray telescope volume from one another will be installed. Due to the low expected signal rate, a window with very low X-ray opacity is needed. Due to a pressure difference of ∝1 bar between detector and the vacuum of CAST this is demanding. The usage of silicon nitride windows is being explored. The current progress of the detector upgrade will be presented.

  10. X-ray micro-beam characterization of a small pixel spectroscopic CdTe detector

    Science.gov (United States)

    Veale, M. C.; Bell, S. J.; Seller, P.; Wilson, M. D.; Kachkanov, V.

    2012-07-01

    A small pixel, spectroscopic, CdTe detector has been developed at the Rutherford Appleton Laboratory (RAL) for X-ray imaging applications. The detector consists of 80 × 80 pixels on a 250 μm pitch with 50 μm inter-pixel spacing. Measurements with an 241Am γ-source demonstrated that 96% of all pixels have a FWHM of better than 1 keV while the majority of the remaining pixels have FWHM of less than 4 keV. Using the Diamond Light Source synchrotron, a 10 μm collimated beam of monochromatic 20 keV X-rays has been used to map the spatial variation in the detector response and the effects of charge sharing corrections on detector efficiency and resolution. The mapping measurements revealed the presence of inclusions in the detector and quantified their effect on the spectroscopic resolution of pixels.

  11. An InGrid based Low Energy X-ray Detector

    CERN Document Server

    Krieger, Christoph; Kaminski, Jochen; Lupberger, Michael; Vafeiadis, Theodoros

    2014-01-01

    An X-ray detector based on the combination of an integrated Micromegas stage with a pixel chip has been built in order to be installed at the CERN Axion Solar Telescope. Due to its high granularity and spatial resolution this detector allows for a topological background suppression along with a detection threshold below $1\\,\\text{keV}$. Tests at the CAST Detector Lab show the detector's ability to detect X-ray photons down to an energy as low as $277\\,\\text{eV}$. The first background data taken after the installation at the CAST experiment underline the detector's performance with an average background rate of $5\\times10^{-5}\\,/\\text{keV}/\\text{cm}^2/\\text{s}$ between 2 and $10\\,\\text{keV}$ when using a lead shielding.

  12. An X-ray scanner prototype based on a novel hybrid gaseous detector

    CERN Document Server

    Iacobaeus, C; Lund-Jensen, B; Peskov, Vladimir

    2007-01-01

    We have developed a prototype of a new type of hybrid X-ray detector. It contains a thin wall (few μm) edge- illuminated lead glass capillary plate (acting as a converter of X-rays photons to primary electrons) combined with a microgap parallel-plate avalanche chamber operating in various gas mixtures at 1 atm. The operation of these converters was studied in a wide range of X-ray energies (from 6 to 60 keV) at incident angles varying from 0° to 90°. The detection efficiency, depending on the geometry, photon's energy, incident angle and the mode of operation, was between a few and 40%. The position resolution achieved was 50 μm in digital form and was practically independent of the photon's energy or gas mixture. The developed detector may open new possibilities for medical imaging, for example in mammography, portal imaging, radiography (including security devices), crystallography and many other applications.

  13. A new cone-beam X-ray CT system with a reduced size planar detector

    International Nuclear Information System (INIS)

    Li Liang; Chen Zhiqiang; Zhang Li; Xing Yuxiang; Kang Kejun

    2006-01-01

    In a traditional cone-beam CT system, the cost of product and computation is very high. The authors propose a transversely truncated cone-beam X-ray CT system with a reduced size detector positioned off-center, in which X-ray beams only cover half of the object. The reduced detector size cuts the cost and the X-ray dose of the CT system. The existing CT reconstruction algorithms are not directly applicable in this new CT system. Hence, the authors develop a BPF-type direct backprojection algorithm. Different from the traditional rebinding methods, our algorithm directly backprojects the pretreated projection data without rebinding. This makes the algorithm compact and computationally more efficient. Finally, some numerical simulations and practical experiments are done to validate the proposed algorithm. (authors)

  14. Performance of low-cost X-ray area detectors with consumer digital cameras

    International Nuclear Information System (INIS)

    Panna, A.; Gomella, A.A.; Harmon, K.J.; Chen, P.; Miao, H.; Bennett, E.E.; Wen, H.

    2015-01-01

    We constructed X-ray detectors using consumer-grade digital cameras coupled to commercial X-ray phosphors. Several detector configurations were tested against the Varian PaxScan 3024M (Varian 3024M) digital flat panel detector. These include consumer cameras (Nikon D800, Nikon D700, and Nikon D3X) coupled to a green emission phosphor in a back-lit, normal incidence geometry, and in a front-lit, oblique incidence geometry. We used the photon transfer method to evaluate detector sensitivity and dark noise, and the edge test method to evaluate their spatial resolution. The essential specifications provided by our evaluation include discrete charge events captured per mm 2 per unit exposure surface dose, dark noise in equivalents of charge events per pixel, and spatial resolution in terms of the full width at half maximum (FWHM) of the detector's line spread function (LSF). Measurements were performed using a tungsten anode X-ray tube at 50 kVp. The results show that the home-built detectors provide better sensitivity and lower noise than the commercial flat panel detector, and some have better spatial resolution. The trade-off is substantially smaller imaging areas. Given their much lower costs, these home-built detectors are attractive options for prototype development of low-dose imaging applications

  15. First examination of CASCADE-X-ray-detector and measurement of neutron-mirrorneutron-oscillation

    International Nuclear Information System (INIS)

    Boehm, B.

    2007-01-01

    The detection of X-radiation is of utmost importance for both fundamental physics and medical diagnostics. This work investigates whether or not the CASCADE detector working principle, first developed for the detection of neutrons, can be adapted for the detection of X-rays. This modular detector concept combines the use of a solid neutron or X-ray converter with the advantages of a counting gas detector. Thus, it gives the possibility to optimize efficiency, dynamics and spatial resolution independently. Firstly, it is necessary to find a suitable converter material that allows for the best possible detector efficiency. In order to do so, a mathematical model of the complete detector system was developed that yields the total efficiency for any given material. Respecting technical constraints, gold and gadolinium showed to be favorable choices. Based on these theoretical considerations a prototype of a CASCADE X-ray detector was built, and measurements for the determination of this detector's efficiency were conducted. In the second part of this work a CASCADE neutron detector was used to conduct the first measurement the neutron-mirrorneutron oscillation time. Mirrormatter was proposed in 1956 by Lee and Yang to allow for symmetry in the description of the universe despite the existence of parity violation. By using neutrons it was possible to determine a lower limit for the oscillation time in this work. (orig.)

  16. Analysis of the charge collection process in solid state X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Kimmel, Nils

    2009-02-12

    Physics with X-rays spans from observing large scales in X-ray astronomy down to small scales in material structure analyses with synchrotron radiation. Both fields of research require imaging detectors featuring spectroscopic resolution for X-rays in an energy range of 0.1 keV to 20.0 keV. Originally driven by the need for an imaging spectrometer on ESA's X-ray astronomy satellite mission XMM-Newton, X-ray pnCCDs were developed at the semiconductor laboratory of the Max-Planck-Institute. The pnCCD is a pixel array detector made of silicon. It is sensitive over a wide band from near infrared- over optical- and UV-radiation up to X-rays. This thesis describes the dynamics of signal electrons from the moment after their generation until their collection in the potential minima of the pixel structure. Experimentally, a pinhole array was used to scan the pnCCD surface with high spatial resolution. Numerical simulations were used as a tool for the modeling of the electrical conditions inside the pnCCD. The results predicted by the simulations were compared with the measurements. Both, experiment and simulation, helped to establish a model for the signal charge dynamics in the energy range from 0.7 keV to 5.5 keV. More generally, the presented work has enhanced the understanding of the detector system on the basis of a physical model. The developed experimental and theoretical methods can be applied to any type of array detector which is based on the full depletion of a semiconductor substrate material. (orig.)

  17. Analysis of the charge collection process in solid state X-ray detectors

    International Nuclear Information System (INIS)

    Kimmel, Nils

    2009-01-01

    Physics with X-rays spans from observing large scales in X-ray astronomy down to small scales in material structure analyses with synchrotron radiation. Both fields of research require imaging detectors featuring spectroscopic resolution for X-rays in an energy range of 0.1 keV to 20.0 keV. Originally driven by the need for an imaging spectrometer on ESA's X-ray astronomy satellite mission XMM-Newton, X-ray pnCCDs were developed at the semiconductor laboratory of the Max-Planck-Institute. The pnCCD is a pixel array detector made of silicon. It is sensitive over a wide band from near infrared- over optical- and UV-radiation up to X-rays. This thesis describes the dynamics of signal electrons from the moment after their generation until their collection in the potential minima of the pixel structure. Experimentally, a pinhole array was used to scan the pnCCD surface with high spatial resolution. Numerical simulations were used as a tool for the modeling of the electrical conditions inside the pnCCD. The results predicted by the simulations were compared with the measurements. Both, experiment and simulation, helped to establish a model for the signal charge dynamics in the energy range from 0.7 keV to 5.5 keV. More generally, the presented work has enhanced the understanding of the detector system on the basis of a physical model. The developed experimental and theoretical methods can be applied to any type of array detector which is based on the full depletion of a semiconductor substrate material. (orig.)

  18. X-ray and gamma ray detector readout system

    Science.gov (United States)

    Tumer, Tumay O; Clajus, Martin; Visser, Gerard

    2010-10-19

    A readout electronics scheme is under development for high resolution, compact PET (positron emission tomography) imagers based on LSO (lutetium ortho-oxysilicate, Lu.sub.2SiO.sub.5) scintillator and avalanche photodiode (APD) arrays. The key is to obtain sufficient timing and energy resolution at a low power level, less than about 30 mW per channel, including all required functions. To this end, a simple leading edge level crossing discriminator is used, in combination with a transimpedance preamplifier. The APD used has a gain of order 1,000, and an output noise current of several pA/ Hz, allowing bipolar technology to be used instead of CMOS, for increased speed and power efficiency. A prototype of the preamplifier and discriminator has been constructed, achieving timing resolution of 1.5 ns FWHM, 2.7 ns full width at one tenth maximum, relative to an LSO/PMT detector, and an energy resolution of 13.6% FWHM at 511 keV, while operating at a power level of 22 mW per channel. Work is in progress towards integration of this preamplifier and discriminator with appropriate coincidence logic and amplitude measurement circuits in an ASIC suitable for a high resolution compact PET instrument. The detector system and/or ASIC can also be used for many other applications for medical to industrial imaging.

  19. A GEANT4 based simulation for pixelated X-ray hybrid detectors

    International Nuclear Information System (INIS)

    Marinho, F.; Akiba, K.

    2015-01-01

    In this letter we present a detailed Monte Carlo approach to simulate pixelated detectors for X-ray applications. It allows us to fully characterize quantities such as interaction probability and reconstructed energy deposits according to beam energy as to evaluate energy and position resolution for comparisons with experimental results. The implementation and use of Monte Carlo truth information is also discussed

  20. Stabilized transistor transformer for self-moving Sirena-1 X-ray flaw detector

    International Nuclear Information System (INIS)

    Krasil'nikov, S.B.; Kristalinskij, A.L.; Lozovoj, L.N.; Markov, S.N.; Sindalovskij, E.I.

    1986-01-01

    Electric circuit of stabilized transistor transformer for self-moving ''Sirena'' type X-ray flaw detector is described. Specifications of the transformer and results of the experimental studies, which can be used when tuning and adjusting the transformer under industrial conditions

  1. The X-ray mirror telescope and the pn-CCD detector of CAST

    CERN Document Server

    Kuster, M; Englhauser, J; Franz, J; Friedrich, P; Hartmann, R; Kang, D; Kotthaus, R; Lutz, Gerhard; Moralez, J; Serber, W; Strüder, L

    2004-01-01

    The Cern Axion Solar Telescope - CAST - uses a prototype 9 Tesla LHC superconducting dipole magnet to search for a hypothetical pseudoscalar particle, the axion, which was proposed by theory in the 1980s to solve the strong CP problem and which could be a dark matter candidate. In CAST a strong magnetic field is used to convert the solar axions to detectable photons via inverse Primakoff effect. The resulting X-rays are thermally distributed in the energy range of 1-7 keV and can be observed with conventional X-ray detectors. The most sensitive detector system of CAST is a pn-CCD detector originally developed for XMM-Newton combined with a Wolter I type X-ray mirror system. The combination of a focusing X-ray optics and a state of the art pn-CCD detector which combines high quantum efficiency, good spacial and energy resolution, and low background improves the sensitivity of the CAST experiment such that for the first time the axion photon coupling constant can be probed beyond the best astrophysical constrai...

  2. X-ray light valve (XLV): a novel detectors' technology for digital mammography

    Science.gov (United States)

    Marcovici, Sorin; Sukhovatkin, Vlad; Oakham, Peter

    2014-03-01

    A novel method, based on X-ray Light Valve (XLV) technology, is proposed for making good image quality yet inexpensive flat panel detectors for digital mammography. The digital mammography markets, particularly in the developing countries, demand quality machines at substantially lower prices than the ones available today. Continuous pressure is applied on x-ray detectors' manufacturers to reduce the flat panel detectors' prices. XLV presents a unique opportunity to achieve the needed price - performance characteristics for direct conversion, x-ray detectors. The XLV based detectors combine the proven, superior, spatial resolution of a-Se with the simplicity and low cost of liquid crystals and optical scanning. The x-ray quanta absorbed by a 200 μm a-Se produce electron - hole pairs that move under an electric field to the top and bottom of a-Se layer. This 2D charge distribution creates at the interface with the liquid crystals a continuous (analog) charge image corresponding to the impinging radiation's information. Under the influence of local electrical charges next to them, the liquid crystals twist proportionally to the charges and vary their light reflectivity. A scanning light source illuminates the liquid crystals while an associated, pixilated photo-detector, having a 42 μm pixel size, captures the light reflected by the liquid crystals and converts it in16 bit words that are transmitted to the machine for image processing and display. The paper will describe a novel XLV, 25 cm x 30 cm, flat panel detector structure and its underlying physics as well as its preliminary performance measured on several engineering prototypes. In particular, the paper will present the results of measuring XLV detectors' DQE, MTF, dynamic range, low contrast resolution and dynamic behavior. Finally, the paper will introduce the new, low cost, XLV detector based, digital mammography machine under development at XLV Diagnostics Inc.

  3. A method to unfold the efficiency of gaseous detectors exposed to broad X-ray spectra

    International Nuclear Information System (INIS)

    Almeida, Gevaldo L. de; Souza, Maria Ines S. de; Lopes, Ricardo T.

    2000-01-01

    A method to obtain the efficiency of a gaseous detector exposed to broad energy X-ray spectra was developed. It consists in the de-convolution of the integrated detector response using the shapes of those spectra as a tool to unfold the aimed detector efficiency curve. For this purpose, the spectra emitted by a X-ray tube under several anode voltages, were properly characterized through measurements with a NaI(Tl) spectrometer. A Lorentz function was then fitted to each of the spectra, and their parameters expressed as a function of the anode voltage, by using polynomial and gaussian fittings. The integral of the product of each Lorentz function, by another unknown Lorentz function, expressing the detector efficiency curve, represents the response of the detector for each anode tension, e.g., each X-ray spectrum. The symbolical integration of that product, produces a general function containing the unknown parameters of the unknown efficiency curve. A non-linear fitting of this general function, to the detector response points, as experimentally obtained, generates the aimed parameters for the efficiency curve. The final detector efficiency curve is obtained after normalization procedures. (author)

  4. Evaluation of K x-ray escape and crosstalk in CdTe detectors and multi-channel detectors

    International Nuclear Information System (INIS)

    Ohtsuchi, Tetsuro; Ohmori, Koichi; Tsutsui, Hiroshi; Baba, Sueki

    1995-01-01

    The simple structure of CdTe semiconductor detectors facilitates their downsizing, and their possible application to radiographic sensors has been studied. The escape of K X-rays from these detectors increases with reduction of their dimensions and affects the measurements of X- and gamma-ray spectra. K X-rays also produce crosstalk in multi-channel detectors with adjacent channels. Therefore, K X-rays which escape from the detector elements degrade both the precision of energy spectra and spatial resolution. The ratios of escape peak integrated counts to total photon counts for various sizes of CdTe single detectors were calculated for gamma rays using the Monte Carlo method. Also, escape and crosstalk ratios were simulated for the CdTe multi-channel detectors. The theoretical results were tested experimentally for 59.54-keV gamma rays from a 241 Am radioactive source. Results showed that escape ratios for single detectors were strongly dependent on element size and thickness. The escape and crosstalk ratios increased with closer channel pitch. The calculated results showed a good agreement with the experimental data. The calculations made it clear that K X-rays which escaped to neighboring channels induced crosstalk more frequently at smaller channel pitch in multichannel detectors. A radiation shielding grid which blocked incident photons between the boundary channels was also tested by experiment and by calculation. It was effective in reducing the probability of escape and crosstalk

  5. A high rate, low noise, x-ray silicon strip detector system

    International Nuclear Information System (INIS)

    Ludewigt, B.; Jaklevic, J.; Kipnis, I.; Rossington, C.; Spieler, H.

    1993-11-01

    An x-ray detector system, based on a silicon strip detector wire-bonded to a low noise charge-senstive amplifier integrated circuit, has been developed for synchrotron radiation experiments which require very high count rates and good energy resolution. Noise measurements and x-ray spectra were taken using a 6 mm long, 55 μm pitch strip detector in conjunction with a prototype 16-channel charge-sensitive preamplifier, both fabricated using standard 1.2 μm CMOS technology. The detector system currently achieves an energy resolution of 350 eV FWHM at 5.9 key, 2 μs peaking time, when cooled to -5 degree C

  6. Automatic analysis of quality of images from X-ray digital flat detectors

    International Nuclear Information System (INIS)

    Le Meur, Y.

    2009-04-01

    Since last decade, medical imaging has grown up with the development of new digital imaging techniques. In the field of X-ray radiography, new detectors replace progressively older techniques, based on film or x-ray intensifiers. These digital detectors offer a higher sensibility and reduced overall dimensions. This work has been prepared with Trixell, the world leading company in flat detectors for medical radiography. It deals with quality control on digital images stemming from these detectors. High quality standards of medical imaging impose a close analysis of the defects that can appear on the images. This work describes a complete process for quality analysis of such images. A particular focus is given on the detection task of the defects, thanks to methods well adapted to our context of spatially correlated defects in noise background. (author)

  7. Slow scan sit detector for x-ray diffraction studies using synchrotron radiation

    International Nuclear Information System (INIS)

    Milch, J.R.

    1978-01-01

    A TV-type x-ray detector using a SIT vidicon has been used for biological diffraction studies at the EMBL outstation at DESY, Hamburg, Germany. The detector converts the two-dimensional diffraction pattern to a charge pattern on the vidicon target, which is read out in the slow-scan mode. This detector has high DOE, no count-rate limit, and is simple and inexpensive to construct. Radiation from the storage ring DORIS was used to study the structure of live muscle at various phases of contraction. Typically the count-rate on the detector was 10 6 x-rays/sec and a total exposure of a few seconds was needed to record the weak diffraction from muscle. This compares with usual exposure times of several hours using a rotating anode generator and film

  8. High energy X-ray photon counting imaging using linear accelerator and silicon strip detectors

    Energy Technology Data Exchange (ETDEWEB)

    Tian, Y., E-mail: cycjty@sophie.q.t.u-tokyo.ac.jp [Department of Bioengineering, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656 (Japan); Shimazoe, K.; Yan, X. [Department of Nuclear Engineering and Management, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656 (Japan); Ueda, O.; Ishikura, T. [Fuji Electric Co., Ltd., Fuji, Hino, Tokyo 191-8502 (Japan); Fujiwara, T. [National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8568 (Japan); Uesaka, M.; Ohno, M. [Nuclear Professional School, the University of Tokyo, 2-22 Shirakata-shirane, Tokai, Ibaraki 319-1188 (Japan); Tomita, H. [Department of Quantum Engineering, Nagoya University, Furo, Chikusa, Nagoya 464-8603 (Japan); Yoshihara, Y. [Department of Nuclear Engineering and Management, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656 (Japan); Takahashi, H. [Department of Bioengineering, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656 (Japan); Department of Nuclear Engineering and Management, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656 (Japan)

    2016-09-11

    A photon counting imaging detector system for high energy X-rays is developed for on-site non-destructive testing of thick objects. One-dimensional silicon strip (1 mm pitch) detectors are stacked to form a two-dimensional edge-on module. Each detector is connected to a 48-channel application specific integrated circuit (ASIC). The threshold-triggered events are recorded by a field programmable gate array based counter in each channel. The detector prototype is tested using 950 kV linear accelerator X-rays. The fast CR shaper (300 ns pulse width) of the ASIC makes it possible to deal with the high instant count rate during the 2 μs beam pulse. The preliminary imaging results of several metal and concrete samples are demonstrated.

  9. High energy X-ray photon counting imaging using linear accelerator and silicon strip detectors

    International Nuclear Information System (INIS)

    Tian, Y.; Shimazoe, K.; Yan, X.; Ueda, O.; Ishikura, T.; Fujiwara, T.; Uesaka, M.; Ohno, M.; Tomita, H.; Yoshihara, Y.; Takahashi, H.

    2016-01-01

    A photon counting imaging detector system for high energy X-rays is developed for on-site non-destructive testing of thick objects. One-dimensional silicon strip (1 mm pitch) detectors are stacked to form a two-dimensional edge-on module. Each detector is connected to a 48-channel application specific integrated circuit (ASIC). The threshold-triggered events are recorded by a field programmable gate array based counter in each channel. The detector prototype is tested using 950 kV linear accelerator X-rays. The fast CR shaper (300 ns pulse width) of the ASIC makes it possible to deal with the high instant count rate during the 2 μs beam pulse. The preliminary imaging results of several metal and concrete samples are demonstrated.

  10. Bio-medical X-ray imaging with spectroscopic pixel detectors

    CERN Document Server

    Butler, A P H; Tipples, R; Cook, N; Watts, R; Meyer, J; Bell, A J; Melzer, T R; Butler, P H

    2008-01-01

    The aim of this study is to review the clinical potential of spectroscopic X-ray detectors and to undertake a feasibility study using a novel detector in a clinical hospital setting. Detectors currently in development, such as Medipix-3, will have multiple energy thresholds allowing for routine use of spectroscopic bio-medical imaging. We have coined the term MARS (Medipix All Resolution System) for bio-medical images that provide spatial, temporal, and energy information. The full clinical significance of spectroscopic X-ray imaging is difficult to predict but insights can be gained by examining both image reconstruction artifacts and the current uses of dual-energy techniques. This paper reviews the known uses of energy information in vascular imaging and mammography, clinically important fields. It then presents initial results from using Medipix-2, to image human tissues within a clinical radiology department. Detectors currently in development, such as Medipix-3, will have multiple energy thresholds allo...

  11. A gas microstrip X-ray detector for soft energy fluorescence EXAFS

    CERN Document Server

    Smith, A D; Derbyshire, G E; Duxbury, D M; Lipp, J; Spill, E J; Stephenson, R

    2001-01-01

    Gas microstrip detectors have been previously developed by the particle physics community, where their robustness, compactness and high counting speed have been recognised. These features are particularly attractive to synchrotron radiation use. In this paper, we describe a gas microstrip detector employing multi-element readout and specifically developed for high count rate fluorescence EXAFS at soft X-ray energies below 4 keV.

  12. Material specific X-ray imaging using an energy-dispersive pixel detector

    Energy Technology Data Exchange (ETDEWEB)

    Egan, Christopher K., E-mail: christopher.egan@manchester.ac.uk [School of Materials, University of Manchester, Manchester M13 9PL (United Kingdom); Wilson, Matthew D.; Veale, Matthew C.; Seller, Paul [STFC Rutherford Appleton Laboratory, Harwell, Didcot, Oxfordshire OX11 0QX (United Kingdom); Jacques, Simon D.M.; Cernik, Robert J. [School of Materials, University of Manchester, Manchester M13 9PL (United Kingdom)

    2014-04-01

    By imaging the X-ray spectral properties or ‘colours’ we have shown how material specific imaging can be performed. Using a pixelated energy-dispersive X-ray detector we record the absorbed and emitted hard X-radiation and measure the energy (colour) and intensity of the photons. Using this technology, we are not only able to obtain attenuation contrast but also to image chemical (elemental) variations inside objects, potentially opening up a very wide range of applications from materials science to medical diagnostics.

  13. A 36-pixel superconducting tunnel junction soft X-ray detector for environmental science applications

    International Nuclear Information System (INIS)

    Friedrich, Stephan; Drury, Owen B.; Cramer, Stephen P.; Green, Peter G.

    2006-01-01

    We are operating a superconducting tunnel junction detector for high-resolution soft X-ray spectroscopy at the Advanced Biological and Environmental X-ray Facility at the Advanced Light Source synchrotron. We have recently upgraded the instrument from 9 to 36 pixels for increased sensitivity. We have also acquired a new digital signal readout to increase the total count rate capabilities to ∼10 6 counts/s while maintaining a high peak-to-background ratio. We report on the performance of the spectrometer, and discuss speciation measurements of chromium in welding aerosols as a typical application of the instrument in environmental science

  14. X-ray inspection of composite materials for aircraft structures using detectors of Medipix type

    International Nuclear Information System (INIS)

    Jandejsek, I; Jakubek, J; Jakubek, M; Krejci, F; Soukup, P; Turecek, D; Vavrik, D; Zemlicka, J; Prucha, P

    2014-01-01

    This work presents an overview of promising X-ray imaging techniques employed for non-destructive defectoscopy inspections of composite materials intended for the Aircraft industry. The major emphasis is placed on non-tomographic imaging techniques which do not require demanding spatial and time measurement conditions. Imaging methods for defects visualisation, delamination detection and porosity measurement of various composite materials such as carbon fibre reinforced polymers and honeycomb sendwiches are proposed. We make use of the new large area WidePix X-ray imaging camera assembled from up to 100 edgeless Medipix type detectors which is highly suitable for this type of measurements

  15. Interference of escape peaks from HPGe(LEPS) detector on characteristic X-ray spectrum

    International Nuclear Information System (INIS)

    Pimjun, S.

    1992-01-01

    Analyses for concentrations of iron, niobium and tantalum in valuable ores by X-ray fluorescence analysis (XRF) method, characteristic X-rays of Fe was interfered by the escape peak of Nb from HPGe(LEPS) detector. To solve this problem, normalizations of Nb spectra with the computer program were used. The result analyzed for Fe, Nb and Ta concentration in the sample, compared to those with the neutron activation analysis (NAA) method, the corresponding errors about 5% were obtained satisfactorily. This method of XRF can be applied efficiently instead of NAA method during the reactor shut-down for a long period

  16. A 36-pixel superconducting tunnel junction soft X-ray detector for environmental science applications

    Energy Technology Data Exchange (ETDEWEB)

    Friedrich, Stephan [Lawrence Livermore National Laboratory, Advanced Detector Group, 7000 East Avenue, L-270, Livermore, CA 94550 (United States) and Lawrence Berkeley National Laboratory, Advanced Biological and Environmental X-ray Facility, 1 Cyclotron Road, Berkeley, CA 94720 (United States)]. E-mail: friedrich1@llnl.gov; Drury, Owen B. [Lawrence Livermore National Laboratory, Advanced Detector Group, 7000 East Avenue, L-270, Livermore, CA 94550 (United States); Lawrence Berkeley National Laboratory, Advanced Biological and Environmental X-ray Facility, 1 Cyclotron Road, Berkeley, CA 94720 (United States); Cramer, Stephen P. [Lawrence Berkeley National Laboratory, Advanced Biological and Environmental X-ray Facility, 1 Cyclotron Road, Berkeley, CA 94720 (United States); Green, Peter G. [University of California Davis, Department of Civil and Environmental Engineering, 1 Shields Avenue, Davis, CA 95616 (United States)

    2006-04-15

    We are operating a superconducting tunnel junction detector for high-resolution soft X-ray spectroscopy at the Advanced Biological and Environmental X-ray Facility at the Advanced Light Source synchrotron. We have recently upgraded the instrument from 9 to 36 pixels for increased sensitivity. We have also acquired a new digital signal readout to increase the total count rate capabilities to {approx}10{sup 6} counts/s while maintaining a high peak-to-background ratio. We report on the performance of the spectrometer, and discuss speciation measurements of chromium in welding aerosols as a typical application of the instrument in environmental science.

  17. A CCD-based area detector for X-ray crystallography using synchrotron and laboratory sources

    International Nuclear Information System (INIS)

    Phillips, W.C.; Li Youli; Stanton, M.; Xie Yuanhui; O'Mara, D.; Kalata, K.

    1993-01-01

    The design and characteristics of a CCD-based area detector suitable for X-ray crystallographic studies using both synchrotron and laboratory sources are described. The active area is 75 mm in diameter, the FWHM of the point response function is 0.20 mm, and for Bragg peaks the dynamic range is 900 and the DQE ∼0.3. The 1320x1035-pixel Kodak CCD is read out into an 8 Mbyte memory system in 0.14 s and digitized to 12 bits. X-ray crystallographic data collected at the NSLS synchrotron from cubic insulin crystals are presented. (orig.)

  18. A new detector system for low energy X-ray fluorescence coupled with soft X-ray microscopy: First tests and characterization

    Energy Technology Data Exchange (ETDEWEB)

    Gianoncelli, Alessandra, E-mail: alessandra.gianoncelli@elettra.eu [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Bufon, Jernej [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); University of Trieste, Piazzale Europa 1, Trieste 34127 (Italy); Ahangarianabhari, Mahdi [Politecnico di Milano, Via Anzani 42, Como 22100 (Italy); INFN Milano, Via Celoria 16, Milano 20133 (Italy); Altissimo, Matteo [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Bellutti, Pierluigi [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Bertuccio, Giuseppe [Politecnico di Milano, Via Anzani 42, Como 22100 (Italy); INFN Milano, Via Celoria 16, Milano 20133 (Italy); Borghes, Roberto [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Carrato, Sergio [University of Trieste, Piazzale Europa 1, Trieste 34127 (Italy); Cautero, Giuseppe [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Fabiani, Sergio [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Giacomini, Gabriele [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Giuressi, Dario [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Kourousias, George [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); Menk, Ralf Hendrik [Elettra-Sincrotrone Trieste S.C.p.A., Strada Statale SS14, km 163.5, Basovizza 34149 (Italy); INFN Trieste, Padriciano 99, Trieste 34149 (Italy); Picciotto, Antonino; Piemonte, Claudio [Fondazione Bruno Kessler, Via Sommarive 18, Trento 38123 (Italy); Rachevski, Alexandre [INFN Trieste, Padriciano 99, Trieste 34149 (Italy); and others

    2016-04-21

    The last decades have witnessed substantial efforts in the development of several detector technologies for X-ray fluorescence (XRF) applications. In spite of the increasing trend towards performing, cost-effective and reliable XRF systems, detectors for soft X-ray spectroscopy still remain a challenge, requiring further study, engineering and customization in order to yield effective and efficient systems. In this paper we report on the development, first characterization and tests of a novel multielement detector system based on low leakage current silicon drift detectors (SDD) coupled to ultra low noise custom CMOS preamplifiers for synchrotron-based low energy XRF. This new system exhibits the potential for improving the count rate by at least an order of magnitude resulting in ten-fold shorter dwell time at an energy resolution similar to that of single element silicon drift detectors.

  19. Large area, low capacitance Si(Li) detectors for high rate x-ray applications

    International Nuclear Information System (INIS)

    Rossington, C.S.; Fine, P.M.; Madden, N.W.

    1992-10-01

    Large area, single-element Si(Li) detectors have been fabricated using a novel geometry which yields detectors with reduced capacitance and hence reduced noise at short amplifier pulse-processing times. A typical device employing the new geometry with a thickness of 6 mm and an active area of 175 mm 2 has a capacitance of only 0.5 pf, compared to 2.9 pf for a conventional planar device with equivalent dimensions. These new low capacitance detectors, used in conjunction with low capacitance field effect transistors, will result in x-ray spectrometers capable of operating at very high count rates while still maintaining excellent energy resolution. The spectral response of the low capacitance detectors to a wide range of x-ray energies at 80 K is comparable to typical state-of-the-art conventional Si(Li) devices. In addition to their low capacitance, the new devices offer other advantages over conventional detectors. Detector fabrication procedures, I-V and C-V characteristics, noise performance, and spectral response to 2-60 keV x-rays are described

  20. Modeling and design of X-rays bidimensional detectors; Modelagem e projeto de detectores bidimensionais para radiacao-X

    Energy Technology Data Exchange (ETDEWEB)

    Quisbert, Elmer Paz Alcon

    2000-03-01

    In this work has been developed the scintillating fiber optic and semiconductor devices based 2-D detector design, modeling and performance evaluation using Monte Carlo methods, for high X-ray energy range (10-140 kV) radiography and tomography applications. These processes allowed us, also, the imaging system parameters and components optimization and appropriate detector design. The model estimated the detectors performance parameters (DQE, MTF and SNR), and radiation risk (in terms of mean absorbed dose in the patient) and to show up how the sequence of physical processes in X-ray detection influence the performance of this imaging PFOC detectors. In this way, the modeling of the detector includes the statistics of the spatial distribution of absorbed X-rays and of X-ray to light conversion, its transmission, and the light quanta conversion into electrons. Also contributions to noise from the detection system chain is included, mainly the CCD detector ambient noise. Performance prediction, based on calculation taken from simulations, illustrates how such detectors meet the exacting requirements of some medical and industrial applications. Also, it is envisaged that our modeling procedure of the imaging system will be suitable not only for investigating how the system components should be best designed but for CT and RD system performance prediction. The powerful techniques would enable us to give advice for future development, in this field, in search of more dose-efficient imaging systems. (author)

  1. CdTe Based Hard X-ray Imager Technology For Space Borne Missions

    Science.gov (United States)

    Limousin, Olivier; Delagnes, E.; Laurent, P.; Lugiez, F.; Gevin, O.; Meuris, A.

    2009-01-01

    CEA Saclay has recently developed an innovative technology for CdTe based Pixelated Hard X-Ray Imagers with high spectral performance and high timing resolution for efficient background rejection when the camera is coupled to an active veto shield. This development has been done in a R&D program supported by CNES (French National Space Agency) and has been optimized towards the Simbol-X mission requirements. In the latter telescope, the hard X-Ray imager is 64 cm² and is equipped with 625µm pitch pixels (16384 independent channels) operating at -40°C in the range of 4 to 80 keV. The camera we demonstrate in this paper consists of a mosaic of 64 independent cameras, divided in 8 independent sectors. Each elementary detection unit, called Caliste, is the hybridization of a 256-pixel Cadmium Telluride (CdTe) detector with full custom front-end electronics into a unique 1 cm² component, juxtaposable on its four sides. Recently, promising results have been obtained from the first micro-camera prototypes called Caliste 64 and will be presented to illustrate the capabilities of the device as well as the expected performance of an instrument based on it. The modular design of Caliste enables to consider extended developments toward IXO type mission, according to its specific scientific requirements.

  2. Photon counting and energy discriminating X-ray detectors. Benefits and applications

    International Nuclear Information System (INIS)

    Walter, David; Zscherpel, Uwe; Ewert, Uwe

    2016-01-01

    Since a few years the direct detection of X-ray photons into electrical signals is possible by usage of highly absorbing photo conducting materials (e.g. CdTe) as detection layer of an underlying CMOS semiconductor X-ray detector. Even NDT energies up to 400 keV are possible today, as well. The image sharpness and absorption efficiency is improved by the replacement of the unsharp scintillation layer (as used at indirect detecting detectors) by a photo conducting layer of much higher thickness. If the read-out speed is high enough (ca. 50 - 100 ns dead time) single X-ray photons can be counted and their energy measured. Read-out noise and dark image correction can be avoided. By setting energy thresholds selected energy ranges of the X-ray spectrum can be detected or suppressed. This allows material discrimination by dual-energy techniques or the reduction of image contributions of scattered radiation, which results in an enhanced contrast sensitivity. To use these advantages in an effective way, a special calibration procedure has to be developed, which considers also time dependent processes in the detection layer. This contribution presents some of these new properties of direct detecting digital detector arrays (DDAs) and shows first results on testing fiber reinforced composites as well as first approaches to dual energy imaging.

  3. Fast photoconductor CdTe detectors for synchrotron x-ray studies

    International Nuclear Information System (INIS)

    Yoo, Sung Shik; Faurie, J.P.; Huang Qiang; Rodricks, B.

    1993-09-01

    The Advanced Photon Source will be that brightest source of synchrotron x-rays when it becomes operational in 1996. During normal operation, the ring will be filled with 20 bunches of positrons with an interbunch spacing of 177 ns and a bunch width of 119 ps. To perform experiments with x-rays generated by positrons on these time scales one needs extremely high speed detectors. To achieve the necessary high speed, we are developing MBE-grown CdTe-base photoconductive position sensitive array detectors. The arrays fabricated have 64 pixels with a gap of 100 μm between pixels. The high speed response of the devices was tested using a short pulse laser. X-ray static measurements were performed using an x-ray tube and synchrotron radiation to study the device's response to flux and wavelength changes. This paper presents the response of the devices to some of these tests and discusses different physics aspects to be considered when designing high speed detectors

  4. CCD [charge-coupled device] sensors in synchrotron x-ray detectors

    International Nuclear Information System (INIS)

    Strauss, M.G.; Naday, I.; Sherman, I.S.; Kraimer, M.R.; Westbrook, E.M.; Zaluzec, N.J.

    1987-01-01

    The intense photon flux from advanced synchrotron light sources, such as the 7-GeV synchrotron being designed at Argonne, require integrating-type detectors. Charge-coupled devices (CCDs) are well suited as synchrotron x-ray detectors. When irradiated indirectly via a phosphor followed by reducing optics, diffraction patterns of 100 cm 2 can be imaged on a 2 cm 2 CCD. With a conversion efficiency of ∼1 CCD electron/x-ray photon, a peak saturation capacity of >10 6 x rays can be obtained. A programmable CCD controller operating at a clock frequency of 20 MHz has been developed. The readout rate is 5 x 10 6 pixels/s and the shift rate in the parallel registers is 10 6 lines/s. The test detector was evaluated in two experiments. In protein crystallography diffraction patterns have been obtained from a lysozyme crystal using a conventional rotating anode x-ray generator. Based on these results we expect to obtain at a synchrotron diffraction images at the rate of ∼1 frame/s or a complete 3-dimensional data set from a single crystal in ∼2 min. 16 refs., 16 figs., 2 tabs

  5. In-line X-ray lensless imaging with lithium fluoride film detectors

    International Nuclear Information System (INIS)

    Bonfigli, F.; Cecilia, A.; Bateni, S. Heidari; Nichelatti, E.; Pelliccia, D.; Somma, F.; Vagovic, P.; Vincenti, M.A.; Baumbach, T.; Montereali, R.M.

    2013-01-01

    In this work, we present preliminary in-line X-ray lensless projection imaging results at a synchrotron facility by using novel solid-state detectors based on non-destructive readout of photoluminescent colour centres in lithium fluoride thin films. The peculiarities of LiF radiation detectors are high spatial resolution on a large field of view, wide dynamic range, versatility and simplicity of use. These properties offered the opportunity to test a broadband X-ray synchrotron source for lensless projection imaging experiments at the TopoTomo beamline of the ANKA synchrotron facility by using a white beam spectrum (3–40 keV). Edge-enhancement effects were observed for the first time on a test object; they are discussed and compared with simulations, on the basis of the colour centre photoluminescence linear response found in the investigated irradiation conditions. -- Highlights: ► We performed broadband X-ray imaging at synchrotron by novel LiF imaging detectors. ► X-ray phase contrast experiments on LiF crystals and thin films were performed. ► Photoluminescent high-quality X-images on a LiF film only 1 μm thick were obtained. ► Edge-enhancement effects were detected and compared with simulations. ► A linearity of colour centre fluorescence response of LiF film was found

  6. Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging.

    Science.gov (United States)

    Iwanczyk, Jan S; Nygård, Einar; Meirav, Oded; Arenson, Jerry; Barber, William C; Hartsough, Neal E; Malakhov, Nail; Wessel, Jan C

    2009-01-01

    The development of an innovative detector technology for photon-counting in X-ray imaging is reported. This new generation of detectors, based on pixellated cadmium telluride (CdTe) and cadmium zinc telluride (CZT) detector arrays electrically connected to application specific integrated circuits (ASICs) for readout, will produce fast and highly efficient photon-counting and energy-dispersive X-ray imaging. There are a number of applications that can greatly benefit from these novel imagers including mammography, planar radiography, and computed tomography (CT). Systems based on this new detector technology can provide compositional analysis of tissue through spectroscopic X-ray imaging, significantly improve overall image quality, and may significantly reduce X-ray dose to the patient. A very high X-ray flux is utilized in many of these applications. For example, CT scanners can produce ~100 Mphotons/mm(2)/s in the unattenuated beam. High flux is required in order to collect sufficient photon statistics in the measurement of the transmitted flux (attenuated beam) during the very short time frame of a CT scan. This high count rate combined with a need for high detection efficiency requires the development of detector structures that can provide a response signal much faster than the transit time of carriers over the whole detector thickness. We have developed CdTe and CZT detector array structures which are 3 mm thick with 16×16 pixels and a 1 mm pixel pitch. These structures, in the two different implementations presented here, utilize either a small pixel effect or a drift phenomenon. An energy resolution of 4.75% at 122 keV has been obtained with a 30 ns peaking time using discrete electronics and a (57)Co source. An output rate of 6×10(6) counts per second per individual pixel has been obtained with our ASIC readout electronics and a clinical CT X-ray tube. Additionally, the first clinical CT images, taken with several of our prototype photon-counting and

  7. Development of an X-ray detector using photodiodes; Desarrollo de un detector de rayos X usando fotodiodos

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez G, J.; Azorin V, J. C.; Sosa A, M. A.; Ceron, P., E-mail: gonzalezgj2012@licifug.ugto.mx [Universidad de Guanajuato, Division de Ciencias e Ingenierias, Loma del Bosque No. 103, Col. Lomas del Campestre, 37150 Leon, Guanajuato (Mexico)

    2016-10-15

    Currently the radiation detectors for medical applications are very high value in the market and are difficult to access as training material. In the Sciences and Engineering Division of the Guanajuato University (Mexico) investigations are carried out related to ionizing radiations, especially with X-rays. To overcome the lack of materials has had to resort to borrowing equipment from other institutions, so its use and availability are intermittent. For these reasons is proposed to design and implement an X-ray detector for the use of the work group and the University. This work aims to build an X-ray semiconductor detector using inexpensive and affordable materials, is also proposed the use of a photodiode sensor and an Arduino analog-digital card and a LCD display showing the data. (Author)

  8. Segmented Monolithic Germanium Detector Arrays for X-ray Absorption Spectroscopy. Final Report

    International Nuclear Information System (INIS)

    Hull, Ethan L.

    2011-01-01

    The experimental results from the Phase I effort were extremely encouraging. During Phase I PHDs Co. made the first strides toward a new detector technology that could have great impact on synchrotron x-ray absorption (XAS) measurements, and x-ray detector technology in general. Detector hardware that allowed critical demonstration measurements of our technology was designed and fabricated. This new technology allows good charge collection from many pixels on a single side of a multi-element monolithic germanium planar detector. The detector technology provides 'dot-like' collection electrodes having very low capacitance. The detector technology appears to perform as anticipated in the Phase I proposal. In particular, the 7-pixel detector studied showed remarkable properties; making it an interesting example of detector physics. The technology is enabled by the use of amorphous germanium contact technology on germanium planar detectors. Because of the scalability associated with the fabrication of these technologies at PHDs Co., we anticipate being able to supply larger detector systems at significantly lower cost than systems made in the conventional manner.

  9. Deconvolving the temporal response of photoelectric x-ray detectors for the diagnosis of pulsed radiations

    International Nuclear Information System (INIS)

    Zou, Shiyang; Song, Peng; Pei, Wenbing; Guo, Liang

    2013-01-01

    Based on the conjugate gradient method, a simple algorithm is presented for deconvolving the temporal response of photoelectric x-ray detectors (XRDs) to reconstruct the resolved time-dependent x-ray fluxes. With this algorithm, we have studied the impact of temporal response of XRD on the radiation diagnosis of hohlraum heated by a short intense laser pulse. It is found that the limiting temporal response of XRD not only postpones the rising edge and peak position of x-ray pulses but also smoothes the possible fluctuations of radiation fluxes. Without a proper consideration of the temporal response of XRD, the measured radiation flux can be largely misinterpreted for radiation pulses of a hohlraum heated by short or shaped laser pulses

  10. First tests of a Medipix-1 pixel detector for X-ray dynamic defectoscopy

    CERN Document Server

    Vavrik, D; Visschers, J; Pospísil, S; Ponchut, C; Zemankova, J

    2002-01-01

    Recent theoretical damage material models describe the dynamic development of voids and microcracks in materials under plastic deformation. For these models, experimental verification is needed. We propose direct and non-destructive observation of the propagation of material damage by measuring changes in transmission of X-rays penetrating a stressed material, using a photon-counting X-ray imager. The present contribution aims to demonstrate the applicability of silicon and gallium-arsenide devices for X-ray transmission measurements with a specimen of high-ductile aluminium alloy under study. The first experiments to determine the resolution and the sensitivity of the proposed method with the Medipix-1 pixel detector are presented.

  11. Dosimetric properties of a radiochromic gel detector for diagnostic X-rays

    International Nuclear Information System (INIS)

    Bero, M.A.

    2007-01-01

    The gel dosimetry method was found to be capable of addressing complicated issues related to dose measurements particularly in modern sophisticated radiotherapy applications. The Ferrous-sulphate Xylenol-orange and Gelatin (FXG) radiochromic gel dosemeter is one of the systems used for such applications. Some chemical dosemeters show different responses for low- and medium-energy X-rays in comparison with high-energy γ-photons. The energy and dose rate dependence of the FXG dose response was examined. In addition to the detector response, other important dosimetric properties of the system were investigated for different X-ray beam qualities with tube voltages in the range 100-300 kV. An orthovoltage X-ray therapy unit was used to irradiate standard sized samples of FXG from different batches for radiation doses in the range 0-20 Gy

  12. Characterization of energy response for photon-counting detectors using x-ray fluorescence

    International Nuclear Information System (INIS)

    Ding, Huanjun; Cho, Hyo-Min; Molloi, Sabee; Barber, William C.; Iwanczyk, Jan S.

    2014-01-01

    Purpose: To investigate the feasibility of characterizing a Si strip photon-counting detector using x-ray fluorescence. Methods: X-ray fluorescence was generated by using a pencil beam from a tungsten anode x-ray tube with 2 mm Al filtration. Spectra were acquired at 90° from the primary beam direction with an energy-resolved photon-counting detector based on an edge illuminated Si strip detector. The distances from the source to target and the target to detector were approximately 19 and 11 cm, respectively. Four different materials, containing silver (Ag), iodine (I), barium (Ba), and gadolinium (Gd), were placed in small plastic containers with a diameter of approximately 0.7 cm for x-ray fluorescence measurements. Linear regression analysis was performed to derive the gain and offset values for the correlation between the measured fluorescence peak center and the known fluorescence energies. The energy resolutions and charge-sharing fractions were also obtained from analytical fittings of the recorded fluorescence spectra. An analytical model, which employed four parameters that can be determined from the fluorescence calibration, was used to estimate the detector response function. Results: Strong fluorescence signals of all four target materials were recorded with the investigated geometry for the Si strip detector. The average gain and offset of all pixels for detector energy calibration were determined to be 6.95 mV/keV and −66.33 mV, respectively. The detector’s energy resolution remained at approximately 2.7 keV for low energies, and increased slightly at 45 keV. The average charge-sharing fraction was estimated to be 36% within the investigated energy range of 20–45 keV. The simulated detector output based on the proposed response function agreed well with the experimental measurement. Conclusions: The performance of a spectral imaging system using energy-resolved photon-counting detectors is very dependent on the energy calibration of the

  13. Development of an X-ray imaging system with SOI pixel detectors

    Energy Technology Data Exchange (ETDEWEB)

    Nishimura, Ryutaro, E-mail: ryunishi@post.kek.jp [School of High Energy Accelerator Science, SOKENDAI (The Graduate University for Advanced Studies), Oho 1-1, Tsukuba, Ibaraki 305-0801 (Japan); Arai, Yasuo; Miyoshi, Toshinobu [Institute of Particle and Nuclear Studies, High Energy Accelerator Research Organization (KEK-IPNS), Oho 1-1, Tsukuba, Ibaraki 305-0801 (Japan); Hirano, Keiichi; Kishimoto, Shunji; Hashimoto, Ryo [Institute of Materials Structure Science, High Energy Accelerator Research Organization (KEK-IMSS), Oho 1-1, Tsukuba, Ibaraki 305-0801 (Japan)

    2016-09-21

    An X-ray imaging system employing pixel sensors in silicon-on-insulator technology is currently under development. The system consists of an SOI pixel detector (INTPIX4) and a DAQ system based on a multi-purpose readout board (SEABAS2). To correct a bottleneck in the total throughput of the DAQ of the first prototype, parallel processing of the data taking and storing processes and a FIFO buffer were implemented for the new DAQ release. Due to these upgrades, the DAQ throughput was improved from 6 Hz (41 Mbps) to 90 Hz (613 Mbps). The first X-ray imaging system with the new DAQ software release was tested using 33.3 keV and 9.5 keV mono X-rays for three-dimensional computerized tomography. The results of these tests are presented. - Highlights: • The X-ray imaging system employing the SOI pixel sensor is currently under development. • The DAQ of the first prototype has the bottleneck in the total throughput. • The new DAQ release solve the bottleneck by parallel processing and FIFO buffer. • The new DAQ release was tested using 33.3 keV and 9.5 keV mono X-rays.

  14. Characterization of spectrometric photon-counting X-ray detectors at different pitches

    Science.gov (United States)

    Jurdit, M.; Brambilla, A.; Moulin, V.; Ouvrier-Buffet, P.; Radisson, P.; Verger, L.

    2017-09-01

    There is growing interest in energy-sensitive photon-counting detectors based on high flux X-ray imaging. Their potential applications include medical imaging, non-destructive testing and security. Innovative detectors of this type will need to count individual photons and sort them into selected energy bins, at several million counts per second and per mm2. Cd(Zn)Te detector grade materials with a thickness of 1.5 to 3 mm and pitches from 800 μm down to 200 μm were assembled onto interposer boards. These devices were tested using in-house-developed full-digital fast readout electronics. The 16-channel demonstrators, with 256 energy bins, were experimentally characterized by determining spectral resolution, count rate, and charge sharing, which becomes challenging at low pitch. Charge sharing correction was found to efficiently correct X-ray spectra up to 40 × 106 incident photons.s-1.mm-2.

  15. Current applications of semiconductor x-ray detectors in chemical analysis

    International Nuclear Information System (INIS)

    Whitehead, N.E.

    1975-07-01

    In the last few years, semiconductor detectors have been used as X-ray detectors with great success, and the routine rapid accumulation of X-ray spectra is now possible. This review surveys the historical development of the detectors, the utilisation, and relative merits of various means of exciting the X-radiation from the elements in the sample, and compares the technique with other methods claiming to offer the capability of simultaneous multi-element analysis. It is concluded that it is of average sensitivity, but offers some advantages from its non-destructive nature, and in some cases its ability to offer information about the spatial distribution of elements in a sample. Other types of analysis may also be possible simultaneously. Sample preparation techniques are reviewed, especially techniques of manufacturing thin samples. An appendix contains details of the very wide variety of samples which have been analysed. More than 350 references are included. (auth.)

  16. Detectors for X-ray diffraction and scattering: current technology and future challenges

    International Nuclear Information System (INIS)

    Bahr, D.; Brugemann, L.; Gerndt, E.

    2003-01-01

    Full text: Detectors are crucial devices determining the quality, the reliability and the throughput of x-ray diffraction (XRD) and scattering investigations. This is of utmost importance in an industrial environment where in many cases untrained personnel or even without human intervention the experiments and data evaluations are running. The currently used technology of 0-dimensional to 2-dim XRD detectors is presented using selected examples. The application specific requirements on e.g. energy range and resolution, count rate limit, background and dynamic range, and size versus price are discussed. Due to the fact that x-ray diffraction investigations are becoming increasingly attractive in science, research and industry the advance in detector technology is pushed beyond existing limits. The discussion of the resultant market opportunities versus the cost of ownership and market entrance barrier is the final section of the presentation

  17. A microprogrammable high-speed data collection system for position sensitive X-ray detectors

    International Nuclear Information System (INIS)

    Hashizume, H.

    1984-01-01

    A high-speed data acquisition system has been designed which collects digital data from one- and two-dimensional position sensitive X-ray detectors at a maximum average data rate of 1 MHz. The system consists of two separate fast buffer memories, a 64 K word by 20-bit main storage, two timers, a display controller, a computer interface and a keyboard, controlled by a specially designed microprogrammable microprocessor. Data collection is performed by executing a microprogram stored in the control storage; data coming from a detector are first accumulated in a small but fast buffer memory by hardware and transferred to the main storage under control of the microprogram. This design not only permits time-resolved data collections but also provides maximum speed, flexibility and cost-effectiveness simultaneously. The system also accepts data from integrated detectors such as TV cameras. The system has been designed for use in experiments at conventional and synchrotron X-ray sources. (orig.)

  18. Discriminating cosmic muons and X-rays based on rise time using a GEM detector

    Science.gov (United States)

    Wu, Hui-Yin; Zhao, Sheng-Ying; Wang, Xiao-Dong; Zhang, Xian-Ming; Qi, Hui-Rong; Zhang, Wei; Wu, Ke-Yan; Hu, Bi-Tao; Zhang, Yi

    2016-08-01

    Gas electron multiplier (GEM) detectors have been used in cosmic muon scattering tomography and neutron imaging over the last decade. In this work, a triple GEM device with an effective readout area of 10 cm × 10 cm is developed, and a method of discriminating between cosmic muons and X-rays based on rise time is tested. The energy resolution of the GEM detector is tested by 55Fe ray source to prove the GEM detector has a good performance. Analysis of the complete signal-cycles allows us to get the rise time and pulse heights. The experiment result indicates that cosmic muons and X-rays can be discriminated with an appropriate rise time threshold. Supported by National Natural Science Foundation of China (11135002, 11275235, 11405077, 11575073)

  19. Development of multiwire gas detectors for X-rays; Desenvolvimento de detectores a gas multifilares para raios-X

    Energy Technology Data Exchange (ETDEWEB)

    Sales, Eraldo de

    2015-06-01

    This work presents the prototype of a 2D position sensitive gas detector for application in X-ray scattering and diffraction experiments. Starting from a detector initially developed for other applications and will show the required changes on the original concept of this device. The strategy used to determine the necessary adaptations were based on searching in the literature for the overall characteristics of a multi-wire X-ray detector (choice of gas, pressure, window, etc.), the use of simulations, implementation of the changes and finally operational tests. Computational tools were used to calculate the mechanical strength and attenuation of the X-ray photons that helped to determine the most appropriate material for the construction of the entrance window. Detector simulations were built with Garfield software and were used to study the overall properties of the detector, and to determine the optimum parameters for the equipment operation. Typical parameters are the distance between the wires, wire diameter, high voltage to be used, among several other parameters. The results obtained showed that the multi-wire detector concept with the implemented adaptations allowed the detection of X-rays. However, depending on the application, it may be necessary improve the resolution of the equipment, in order to have a better description of the collected data. Several ideas are suggested for this improvement. It is also presented interesting results obtained with a microscopic pattern detector called triple GEM. This device belongs to the Gas Detectors Development group (GDD group) at CERN and was used in my training at this institution. The results showed the potential of the equipment for detection of X-rays. The results and simulations presented in this work, confirmed that the changes in the concept of the original detector permitted it use on X-ray detection applications. Also, it was possible to obtain several indications for further optimization, which may

  20. X-ray Imaging Using a Hybrid Photon Counting GaAs Pixel Detector

    CERN Document Server

    Schwarz, C; Göppert, R; Heijne, Erik H M; Ludwig, J; Meddeler, G; Mikulec, B; Pernigotti, E; Rogalla, M; Runge, K; Smith, K M; Snoeys, W; Söldner-Rembold, S; Watt, J

    1999-01-01

    The performance of hybrid GaAs pixel detectors as X-ray imaging sensors were investigated at room temperature. These hybrids consist of 300 mu-m thick GaAs pixel detectors, flip-chip bonded to a CMOS Single Photon Counting Chip (PCC). This chip consists of a matrix of 64 x 64 identical square pixels (170 mu-m x 170 mu-m) and covers a total area of 1.2 cm**2. The electronics in each cell comprises a preamplifier, a discriminator with a 3-bit threshold adjust and a 15-bit counter. The detector is realized by an array of Schottky diodes processed on semi-insulating LEC-GaAs bulk material. An IV-charcteristic and a detector bias voltage scan showed that the detector can be operated with voltages around 200 V. Images of various objects were taken by using a standard X-ray tube for dental diagnostics. The signal to noise ratio (SNR) was also determined. The applications of these imaging systems range from medical applications like digital mammography or dental X-ray diagnostics to non destructive material testing (...

  1. Evaluation of a hybrid photon counting pixel detector for X-ray polarimetry

    International Nuclear Information System (INIS)

    Michel, T.; Durst, J.

    2008-01-01

    It has already been shown in literature that X-ray sensitive CCDs can be used to measure the degree of linear polarization of X-rays using the effect that photoelectrons are emitted with a non-isotropic angular distribution in respect to the orientation of the electric field vector of impinging photons. Up to now hybrid semiconductor pixel detectors like the Timepix-detector have never been used for X-ray polarimetry. The main reason for this is that the pixel pitch is large compared to CCDs which results in a much smaller analyzing power. On the other hand, the active thickness of the sensor layer can be larger than in CCDs leading to an increased efficiency. Therefore hybrid photon counting pixel detectors may be used for imaging and polarimetry at higher photon energies. For irradiation with polarized X-ray photons we were able to measure an asymmetry between vertical and horizontal double hit events in neighboring pixels of the hybrid photon counting Timepix-detector at room temperature. For the specific spectrum used in our experiment an average polarization asymmetry of (0.96±0.02)% was measured. Additionally, the Timepix-detector with its spectroscopic time-over-threshold-mode was used to measure the dependence of the polarization asymmetry on energy deposition in the detector. Polarization asymmetries between 0.2% at 29 keV and 3.4% at 78 keV energy deposition were determined. The results can be reproduced with our EGS4-based Monte-Carlo simulation

  2. Cascaded systems analysis of charge sharing in cadmium telluride photon-counting x-ray detectors.

    Science.gov (United States)

    Tanguay, Jesse; Cunningham, Ian A

    2018-05-01

    Single-photon-counting (SPC) and spectroscopic x-ray detectors are under development in academic and industry laboratories for medical imaging applications. The spatial resolution of SPC and spectroscopic x-ray detectors is an important design criterion. The purpose of this article was to extend the cascaded systems approach to include a description of the spatial resolution of SPC and spectroscopic x-ray imaging detectors. A cascaded systems approach was used to model reabsorption of characteristic x rays, Coulomb repulsion, and diffusion in SPC and spectroscopic x-ray detectors. In addition to reabsorption, diffusion, and Coulomb repulsion, the model accounted for x-ray conversion to electron-hole (e-h) pairs, integration of e-h pairs in detector elements, electronic noise, and energy thresholding. The probability density function (PDF) describing the number of e-h pairs was propagated through each stage of the model and was used to derive new theoretical expressions for the large-area gain and modulation transfer function (MTF) of CdTe SPC x-ray detectors, and the energy bin sensitivity functions and MTFs of CdTe spectroscopic detectors. Theoretical predictions were compared with the results of MATLAB-based Monte Carlo (MC) simulations and published data. Comparisons were also made with the MTF of energy-integrating systems. Under general radiographic conditions, reabsorption, diffusion, and Coulomb repulsion together artificially inflate count rates by 20% to 50%. For thicker converters (e.g. 1000 μm) and larger detector elements (e.g. 500 μm pixel pitch) these processes result in modest inflation (i.e. ∼10%) in apparent count rates. Our theoretical and MC analyses predict that SPC MTFs will be degraded relative to those of energy-integrating systems for fluoroscopic, general radiographic, and CT imaging conditions. In most cases, this degradation is modest (i.e., ∼10% at the Nyquist frequency). However, for thicker converters, the SPC MTF can be degraded

  3. Operation of an InGrid based X-ray detector at the CAST experiment

    Science.gov (United States)

    Krieger, Christoph; Desch, Klaus; Kaminski, Jochen; Lupberger, Michael

    2018-02-01

    The CERN Axion Solar Telescope (CAST) is searching for axions and other particles which could be candidates for DarkMatter and even Dark Energy. These particles could be produced in the Sun and detected by a conversion into soft X-ray photons inside a strong magnetic field. In order to increase the sensitivity for physics beyond the Standard Model, detectors with a threshold below 1 keV as well as efficient background rejection methods are required to compensate for low energies and weak couplings resulting in very low detection rates. Those criteria are fulfilled by a detector utilizing the combination of a pixelized readout chip with an integrated Micromegas stage. These InGrid (Integrated Grid) devices can be build by photolithographic postprocessing techniques, resulting in a close to perfect match of grid and pixels facilitating the detection of single electrons on the chip surface. The high spatial resolution allows for energy determination by simple electron counting as well as for an event-shape based analysis as background rejection method. Tests at an X-ray generator revealed the energy threshold of an InGrid based X-ray detector to be well below the carbon Kα line at 277 eV. After the successful demonstration of the detectors key features, the detector was mounted at one of CAST's four detector stations behind an X-ray telescope in 2014. After several months of successful operation without any detector related interruptions, the InGrid based X-ray detector continues data taking at CAST in 2015. During operation at the experiment, background rates in the order of 10-5 keV-1 cm-2 s-1 have been achieved by application of a likelihood based method discriminating the non-photon background originating mostly from cosmic rays. For continued operation in 2016, an upgraded InGrid based detector is to be installed among other improvements including decoupling and sampling of the signal induced on the grid as well as a veto scintillator to further lower the

  4. Trends in hard X-ray fluorescence mapping: environmental applications in the age of fast detectors

    Energy Technology Data Exchange (ETDEWEB)

    Lombi, E.; Donner, E. [University of South Australia, Centre for Environmental Risk Assessment and Remediation, Mawson Lakes, South Australia (Australia); CRC CARE, PO Box 486, Salisbury, South Australia (Australia); Jonge, M.D. de; Paterson, D. [Australian Synchrotron, X-ray Fluorescence Microscopy, 800 Blackburn Road, Clayton, Victoria (Australia); Ryan, C.G. [CSIRO Earth Science and Resource Engineering, Normanby Road, Clayton, Victoria (Australia)

    2011-06-15

    Environmental samples are extremely diverse but share a tendency for heterogeneity and complexity. This heterogeneity poses methodological challenges when investigating biogeochemical processes. In recent years, the development of analytical tools capable of probing element distribution and speciation at the microscale have allowed this challenge to be addressed. Of these available tools, laterally resolved synchrotron techniques such as X-ray fluorescence mapping are key methods for the in situ investigation of micronutrients and inorganic contaminants in environmental samples. This article demonstrates how recent advances in X-ray fluorescence detector technology are bringing new possibilities to environmental research. Fast detectors are helping to circumvent major issues such as X-ray beam damage of hydrated samples, as dwell times during scanning are reduced. They are also helping to reduce temporal beamtime requirements, making particularly time-consuming techniques such as micro X-ray fluorescence ({mu}XRF) tomography increasingly feasible. This article focuses on {mu}XRF mapping of nutrients and metalloids in environmental samples, and suggests that the current divide between mapping and speciation techniques will be increasingly blurred by the development of combined approaches. (orig.)

  5. Wide Bandgap Semiconductor Detector Optimization for Flash X-Ray Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Roecker, Caleb Daniel [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Schirato, Richard C. [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-11-17

    Charge trapping, resulting in a decreased and spatially dependent electric field, has long been a concern for wide bandgap semiconductor detectors. While significant work has been performed to characterize this degradation at varying temperatures and radiation environments, this work concentrates upon examining the event-to-event response in a flash X-ray environment. The following work investigates if charge trapping is a problem for CZT detectors, with particular emphasis on flash X-ray radiation fields at cold temperatures. Results are compared to a non-flash radiation field, using an Am-241 alpha source and similar temperature transitions. Our ability to determine if a response change occurred was hampered by the repeatability of our flash X-ray systems; a small response change was observed with the Am-241 source. Due to contrast of these results, we are in the process of revisiting the Am-241 measurements in the presence of a high radiation environment. If the response change is more pronounced in the high radiation environment, a similar test will be performed in the flash X-ray environment.

  6. Quantitative microanalysis in the analytical electronmicroscope using an HPGe-x ray detector

    International Nuclear Information System (INIS)

    Grogger, W.

    1994-01-01

    Energy dispersive x-ray spectrometry (EDX) is a routine method for determining the chemical composition of a sample in the analytical electronmicroscope. Since some years high purity germanium x-ray detectors (HPGe) are commercially available for use in EDX. This new type of detector offers some advantages over the commonly used Si (Li) detector: better energy resolution, better detector efficiency for high energy lines (> 30 keV) and better stability against exterior influences. For quantitative analysis one needs sensitivity factors (k-factors), which correlate the measured intensity to the concentration of a specific element. These k-factors can be calculated or determined experimentally. For a precise quantitative analysis of light elements measured k-factors are absolutely necessary. In this study k-factors were measured with an HPGe detector using standards. The accuracy of the k-factors was proved using some examples of practical relevance. Additionally some special features of the HPGe detector were examined, which lead to a better understanding of EDX spectrometry using an HPGe detector (escape lines, icing of the detector, artifacts). (author)

  7. Development and features of an X-ray detector with high spatial resolution

    International Nuclear Information System (INIS)

    Hartmann, H.

    1979-09-01

    A laboratory model of an X-ray detector with high spatial resolution was developed and constructed. It has no spectral resolution, but a local resolution of 20 μm which is about ten times as high as that of position-sensitive proportional counters and satisfies the requirements of the very best Wolter telescopes with regard to spatial resolution. The detector will be used for laboratory tests of the 80 cm Wolter telescope which is being developed for Spacelab flights. The theory of the wire grid detector and the physics of the photoelectric effect has been developed, and model calculations and numerical calculations have been carried out. (orig./WB) [de

  8. Performance study of monochromatic synchrotron X-ray computed tomography using a linear array detector

    Energy Technology Data Exchange (ETDEWEB)

    Kazama, Masahiro; Takeda, Tohoru; Itai, Yuji [Tsukuba Univ., Ibaraki (Japan). Inst. of Clinical Medicine; Akiba, Masahiro; Yuasa, Tetsuya; Hyodo, Kazuyuki; Ando, Masami; Akatsuka, Takao

    1997-09-01

    Monochromatic x-ray computed tomography (CT) using synchrotron radiation (SR) is being developed for detection of non-radioactive contrast materials at low concentration for application in clinical diagnosis. A new SR-CT system with improved contrast resolution, was constructed using a linear array detector which provides wide dynamic ranges and a double monochromator. The performance of this system was evaluated in a phantom and a rat model of brain ischemia. This system consists of a silicon (111) double crystal monochromator, an x-ray shutter, an ionization chamber, x-ray slits, a scanning table for the target organ, and an x-ray linear array detector. The research was carried out at the BLNE-5A bending magnet beam line of the Tristan Accumulation Ring in KEK, Japan. In this experiment, the reconstructed image of the spatial-resolution phantom clearly showed the 1 mm holes. At 1 mm slice thickness, the above K-edge image of the phantom showed contrast resolution at the concentration of 200 {mu}g/ml iodine-based contrast materials whereas the K-edge energy subtraction image showed contrast resolution at the concentration of 500 {mu}g/ml contrast materials. The cerebral arteries filled with iodine microspheres were clearly revealed, and the ischemic regions at the right temporal lobe and frontal lobe were depicted as non-vascular regions. The measured minimal detectable concentration of iodine on the above K-edge image is about 6 times higher than the expected value of 35.3 {mu}g/ml because of the high dark current of this detector. Thus, the use of a CCD detector which is cooled by liquid nitrogen to improve the dynamic range of the detector, is being under construction. (author)

  9. Nanosecond X-ray detector based on high resistivity ZnO single crystal semiconductor

    Energy Technology Data Exchange (ETDEWEB)

    Zhao, Xiaolong; He, Yongning, E-mail: yongning@mail.xjtu.edu.cn; Peng, Wenbo; Huang, Zhiyong; Qi, Xiaomeng; Pan, Zijian; Zhang, Wenting [School of Electronic and Information Engineering, Xi' an Jiaotong University, Xi' an 710049 (China); Chen, Liang; Liu, Jinliang; Zhang, Zhongbing; Ouyang, Xiaoping [Radiation Detection Research Center, Northwest Institute of Nuclear Technology, Xi' an 710024 (China)

    2016-04-25

    The pulse radiation detectors are sorely needed in the fields of nuclear reaction monitoring, material analysis, astronomy study, spacecraft navigation, and space communication. In this work, we demonstrate a nanosecond X-ray detector based on ZnO single crystal semiconductor, which emerges as a promising compound-semiconductor radiation detection material for its high radiation tolerance and advanced large-size bulk crystal growth technique. The resistivity of the ZnO single crystal is as high as 10{sup 13} Ω cm due to the compensation of the donor defects (V{sub O}) and acceptor defects (V{sub Zn} and O{sub i}) after high temperature annealing in oxygen. The photoconductive X-ray detector was fabricated using the high resistivity ZnO single crystal. The rise time and fall time of the detector to a 10 ps pulse electron beam are 0.8 ns and 3.3 ns, respectively, indicating great potential for ultrafast X-ray detection applications.

  10. X-ray sky

    International Nuclear Information System (INIS)

    Gruen, M.; Koubsky, P.

    1977-01-01

    The history is described of the discoveries of X-ray sources in the sky. The individual X-ray detectors are described in more detail, i.e., gas counters, scintillation detectors, semiconductor detectors, and the principles of X-ray spectrometry and of radiation collimation aimed at increased resolution are discussed. Currently, over 200 celestial X-ray sources are known. Some were identified as nebulae, in some pulsations were found or the source was identified as a binary star. X-ray bursts of novae were also observed. The X-ray radiation is briefly mentioned of spherical star clusters and of extragalactic X-ray sources. (Oy)

  11. Simulated and experimental spectroscopic performance of GaAs X-ray pixel detectors

    International Nuclear Information System (INIS)

    Bisogni, M.G.; Cola, A.; Fantacci, M.E.

    2001-01-01

    In pixel detectors, the electrode geometry affects the signal shape and therefore the spectroscopic performance of the device. This effect is enhanced in semiconductors where carrier trapping is relevant. In particular, semi insulating (SI) GaAs crystals present an incomplete charge collection due to a high concentration of deep traps in the bulk. In the last few years, SI GaAs pixel detectors have been developed as soft X-ray detectors for medical imaging applications. In this paper, we present a numerical method to evaluate the local charge collection properties of pixel detectors. A bi-dimensional description has been used to represent the detector geometry. According to recent models, the active region of a reverse biased SI GaAs detector is almost neutral. Therefore, the electrostatic potential inside a full active detector has been evaluated using the Laplace equation. A finite difference method with a fixed step orthogonal mesh has been adopted. The photon interaction point has been generated with a Monte Carlo method according to the attenuation length of a monochromatic X-ray beam in GaAs. The number of photogenerated carriers for each interaction has been extracted using a gaussian distribution. The induced signal on the collecting electrode has been calculated according to the Ramo's theorem and the trapping effect has been modeled introducing electron and hole lifetimes. The noise of the charge preamplifier have been also taken into account. A comparison between simulated and experimental X-ray spectra from a 241 Am source acquired with different GaAs pixel detectors has been carried out

  12. The simulation of charge sharing in semiconductor X-ray pixel detectors

    CERN Document Server

    Mathieson, K; O'Shea, V; Passmore, M S; Rahman, M; Smith, K M; Watt, J; Whitehill, C

    2002-01-01

    Two simulation packages were used to model the sharing of charge, due to the scattering and diffusion of carriers, between adjacent pixel elements in semiconductors X-ray detectors. The X-ray interaction and the consequent multiple scattering was modelled with the aid of the Monte Carlo package, MCNP. The resultant deposited charge distribution was then used to create the charge cloud profile in the finite element semiconductor simulation code MEDICI. The analysis of the current pulses induced on pixel electrodes for varying photon energies was performed for a GaAs pixel detector. For a pixel pitch of 25 mu m, the charge lost to a neighbouring pixel was observed to be constant, at 0.6%, through the energies simulated. Ultimately, a fundamental limit on the pixel element size for imaging and spectroscopic devices may be set due to these key physical principles.

  13. Performance of room temperature mercuric iodide (HgI2) detectors in the ultra low energy x-ray region

    International Nuclear Information System (INIS)

    Dabrowski, A.J.; Iwanczyk, J.S.; Barton, J.B.; Huth, G.C.; Whited, R.; Ortale, C.; Economou, T.E.; Turkevich, A.L.

    1980-01-01

    Performance of room temperature mercuric iodide x-ray spectrometers has been recently improved through new fabrication techniques and further development of low noise associated electronic systems. This progress has extended the range of measurements to the ultra low energy x-ray region at room temperature. This paper reports the study of the effect of contact material on the performance of HgI 2 detectors in the low energy x-ray region

  14. Performance characteristics needed for protein crystal diffraction x-ray detectors

    International Nuclear Information System (INIS)

    Westbrook, E. M.

    1999-01-01

    During the 1990's, macromolecular crystallography became progressively more dependent on synchrotrons X-ray sources for diffraction data collection. Detectors of this diffraction data at synchrotrons beamlines have evolved over the decade, from film to image phosphor plates, and then to CCD systems. These changes have been driven by the data quality and quantity improvements each newer detector technology provided. The improvements have been significant. It is likely that newer detector technologies will be adopted at synchrotron beamlines for crystallographic diffraction data collection in the future, but these technologies will have to compete with existing CCD detector systems which are already excellent and are getting incrementally better in terms of size, speed, efficiency, and resolving power. Detector development for this application at synchrotrons must concentrate on making systems which are bigger and faster than CCDs and which can capture weak data more efficiently. And there is a need for excellent detectors which are less expensive than CCD systems

  15. Transmission properties of barite mortar using X-ray spectra measured with Cd Te detector

    Energy Technology Data Exchange (ETDEWEB)

    Santos, J. C.; Mariano, L.; Costa, P. R. [Universidade de Sao Paulo, Instituto de Fisica, Rua do Matao Travessa R. 187, Cidade Universitaria, 05508-090 Sao Paulo (Brazil); Tomal, A., E-mail: josilene@usp.br [Universidade Federal de Goias, Instituto de Fisica, Campus Samambaia, 74001-970 Goiania (Brazil)

    2014-08-15

    Current methods for calculating X-ray shielding barriers do not take into account spectral distribution of the beam transmitted by the protective material. This consideration is important in dose estimations for radiation workers and general public in diagnostic radiology facilities. The aim of the present study was to estimate barite mortar attenuation curves using X-ray spectra weighted by a workload distribution. These curves were described in units of ambient dose equivalent (H (10)), since it is the radiation quantity adopted by IAEA for dose assessment in medical environment. Attenuation curves were determined using the optimized model for shielding evaluation presented by Costa and Caldas (2002). Workload distribution presented by Simpkin (1996), measured primary spectra and mass attenuation coefficients of barite mortar were used as input data in this model. X-ray beams in diagnostic energy range were generated by an industrial X-ray tube with 3 mm of aluminum additional filtration. Primary experimental spectra were measured by a Cd Te detector and corrected by the response function of detector by means of a stripping procedure. Air kerma measurements were performed using an ionization chamber for normalization purpose of the spectra. The corrected spectra presented good agreement with spectra generated by a semi-empirical model. The variation of the ambient dose equivalent as a function of barite mortar thickness was calculated. Using these data, it was estimated the optimized thickness of protective barrier needed for shielding a particular area in an X-ray imaging facility. The results obtained for primary protective barriers exhibit qualitative agreement with those presented in literature. (Author)

  16. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    Energy Technology Data Exchange (ETDEWEB)

    Wilson, Matthew D., E-mail: Matt.Wilson@stfc.ac.uk; Seller, Paul; Veale, Matthew C. [Science and Technology Facilities Council, Rutherford Appleton Laboratory, Harwell Campus,UK (United Kingdom); Connolley, Thomas [Diamond Light Source, I12 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal [Diamond Light Source, B16 Beamline, Harwell Campus, Didcot, Oxfordshire (United Kingdom); Grant, Patrick S.; Liotti, Enzo; Lui, Andrew [Department of Materials, University of Oxford Parks Road, Oxford (United Kingdom)

    2016-07-27

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm{sup 2} with one of the 80×80 pixels imaging an area equivalent to 13µm{sup 2}. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  17. An X-Ray facility to perform irradiation tests and TID studies on electronics and detectors

    CERN Document Server

    Brundu, Davide; Cadeddu, Sandro; Wyllie, Ken; Ciambrone, Paolo

    2018-01-01

    The X-Ray irradiation system of the LHCb group, installed in Cagliari, is presented; with a particular focus on the setup configuration and dose rate calibration. The system can be used to perform Total Ionizing Dose (TID) studies for detectors, readout and front-end electronics. It was already used to test the nSYNC chip, an ASIC for the readout of the LHCb upgraded muon system.

  18. Design of soft-X-ray tomographic system in WEST using GEM detectors

    Czech Academy of Sciences Publication Activity Database

    Mazon, D.; Chernyshova, M.; Jiolat, G.; Czarski, T.; Malard, P.; Kowalska-Strzeciwilk, E.; Jablonski, S.; Figacz, W.; Zagorski, R.; Kubkowska, M.; Kasprowicz, G.; Pozniak, K.; Zabolotny, W.; Larroque, S.; Verger, J.-M.; O’Mullane, M.; Mlynář, Jan; Byszuk, A.; Wojenski, A.

    96-97, October (2015), s. 856-860 ISSN 0920-3796. [Symposium on Fusion Technology 2014(SOFT-28)/28./. San Sebastián, 29.09.2014-03.10.2014] Institutional support: RVO:61389021 Keywords : Soft-X-ray * Gas detector * Tomography * WEST * Tokamak Subject RIV: JF - Nuclear Energetics OBOR OECD: Nuclear related engineering Impact factor: 1.301, year: 2015 http://www.sciencedirect.com/science/article/pii/S0920379615002215

  19. X-ray inspection of composite materials for aircraft structures using detectors of Medipix type

    Czech Academy of Sciences Publication Activity Database

    Jandejsek, Ivan; Jakůbek, J.; Jakůbek, M.; Průcha, P.; Krejčí, F.; Soukup, P.; Tureček, D.; Vavřík, Daniel; Žemlička, J.

    2014-01-01

    Roč. 9, May (2014), C05062 ISSN 1748-0221. [International workshop on radiation imaging detectors /15./. Paris, 23.06.2013-27.06.2013] Institutional support: RVO:68378297 Keywords : inspection with x- rays * detection of defects Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering Impact factor: 1.399, year: 2014 http://iopscience.iop.org/1748-0221/9/05/C05062

  20. Energy dispersive detector for white beam synchrotron x-ray fluorescence imaging

    International Nuclear Information System (INIS)

    Wilson, Matthew D.; Seller, Paul; Veale, Matthew C.; Connolley, Thomas; Dolbnya, Igor P.; Malandain, Andrew; Sawhney, Kawal; Grant, Patrick S.; Liotti, Enzo; Lui, Andrew

    2016-01-01

    A novel, “single-shot” fluorescence imaging technique has been demonstrated on the B16 beamline at the Diamond Light Source synchrotron using the HEXITEC energy dispersive imaging detector. A custom made furnace with 200µm thick metal alloy samples was positioned in a white X-ray beam with a hole made in the furnace walls to allow the transmitted beam to be imaged with a conventional X-ray imaging camera consisting of a 500 µm thick single crystal LYSO scintillator, mirror and lens coupled to an AVT Manta G125B CCD sensor. The samples were positioned 45° to the incident beam to enable simultaneous transmission and fluorescence imaging. The HEXITEC detector was positioned at 90° to the sample with a 50 µm pinhole 13 cm from the sample and the detector positioned 2.3m from pinhole. The geometric magnification provided a field of view of 1.1×1.1mm"2 with one of the 80×80 pixels imaging an area equivalent to 13µm"2. Al-Cu alloys doped with Zr, Ag and Mo were imaged in transmission and fluorescence mode. The fluorescence images showed that the dopant metals could be simultaneously imaged with sufficient counts on all 80x80 pixels within 60 s, with the X-ray flux limiting the fluorescence imaging rate. This technique demonstrated that it is possible to simultaneously image and identify multiple elements on a spatial resolution scale ~10µm or higher without the time consuming need to scan monochromatic energies or raster scan a focused beam of X-rays. Moving to high flux beamlines and using an array of detectors could improve the imaging speed of the technique with element specific imaging estimated to be on a 1 s timescale.

  1. A study on characteristics of X-ray detector for CCD-based EPID

    International Nuclear Information System (INIS)

    Chung, Yong Hyun

    1999-02-01

    The combination of the metal plate/phosphor screen as a x-ray detector with a CCD camera is the most popular detector system among various electronic portal imaging devices (EPIDs). There is a need to optimize the thickness of the metal plate/phosphor screen with high detection efficiency and high spatial resolution for effective transferring of anatomical information. In this study, the thickness dependency on the detection efficiency and the spatial resolution of the metal plate/phosphor screen was investigated by calculation and measurement. The result can be used to determine the optimal thickness of the metal plate as well as of the phosphor screen for the x-ray detector design of therapeutic x-ray imaging and for any specific application. Bremsstrahlung spectrum was calculated by Monte Carlo simulation and by Schiff formula. The detection efficiency was calculated from the total absorbed energy in the phosphor screen using the Monte Carlo simulation and the light output was measured. The spatial resolution, which was defined from the spatial distribution of the absorbed energy, was also calculated and the edge spread function was measured. It was found that the detection efficiency and the spatial resolution were mainly determined by the thickness of metal plate and phosphor screen, respectively. It was also revealed that the detection efficiency and the spatial resolution have trade-off in term of the thickness of the phosphor screen. As the phosphor thickness increases, the detection efficiency increases but the spatial resolution decreases. The curve illustrating the trade-off between the detection efficiency and the spatial resolution of the metal plate/phosphor screen detector is obtained as a function of the phosphor thickness. Based on the calculations, prototype CCD-based EPID was developed and then tested by acquiring phantom images for 6 MV x-ray beam. While, among the captured images, each frame suffered from quantum noise, the frame averaging

  2. Advancing the Technology of Monolithic CMOS detectors for their use as X-ray Imaging Spectrometers

    Science.gov (United States)

    Kenter, Almus

    The Smithsonian Astrophysical Observatory (SAO) proposes a two year program to further advance the scientific capabilities of monolithic CMOS detectors for use as x-ray imaging spectrometers. This proposal will build upon the progress achieved with funding from a previous APRA proposal that ended in 2013. As part of that previous proposal, x- ray optimized, highly versatile, monolithic CMOS imaging detectors and technology were developed and tested. The performance and capabilities of these devices were then demonstrated, with an emphasis on the performance advantages these devices have over CCDs and other technologies. The developed SAO/SRI-Sarnoff CMOS devices incorporate: Low noise, high sensitivity ("gain") pixels; Highly parallel on-chip signal chains; Standard and very high resistivity (30,000Ohm-cm) Si; Back-Side thinning and passivation. SAO demonstrated the performance benefits of each of these features in these devices. This new proposal high-lights the performance of this previous generation of devices, and segues into new technology and capability. The high sensitivity ( 135uV/e) 6 Transistor (6T) Pinned Photo Diode (PPD) pixels provided a large charge to voltage conversion gain to the detect and resolve even small numbers of photo electrons produced by x-rays. The on-chip, parallel signal chain processed an entire row of pixels in the same time that a CCD requires to processes a single pixel. The resulting high speed operation ( 1000 times faster than CCD) provide temporal resolution while mitigating dark current and allowed room temperature operation. The high resistivity Si provided full (over) depletion for thicker devices which increased QE for higher energy x-rays. In this proposal, SAO will investigate existing NMOS and existing PMOS devices as xray imaging spectrometers. Conventional CMOS imagers are NMOS. NMOS devices collect and measure photo-electrons. In contrast, PMOS devices collect and measure photo-holes. PMOS devices have various

  3. Highly multiplexible thermal kinetic inductance detectors for x-ray imaging spectroscopy

    International Nuclear Information System (INIS)

    Ulbricht, Gerhard; Mazin, Benjamin A.; Szypryt, Paul; Walter, Alex B.; Bockstiegel, Clint; Bumble, Bruce

    2015-01-01

    For X-ray imaging spectroscopy, high spatial resolution over a large field of view is often as important as high energy resolution, but current X-ray detectors do not provide both in the same device. Thermal Kinetic Inductance Detectors (TKIDs) are being developed as they offer a feasible way to combine the energy resolution of transition edge sensors with pixel counts approaching CCDs and thus promise significant improvements for many X-ray spectroscopy applications. TKIDs are a variation of Microwave Kinetic Inductance Detectors (MKIDs) and share their multiplexibility: working MKID arrays with 2024 pixels have recently been demonstrated and much bigger arrays are under development. In this work, we present a TKID prototype, which is able to achieve an energy resolution of 75 eV at 5.9 keV, even though its general design still has to be optimized. We further describe TKID fabrication, characterization, multiplexing, and working principle and demonstrate the necessity of a data fitting algorithm in order to extract photon energies. With further design optimizations, we expect to be able to improve our TKID energy resolution to less than 10 eV at 5.9 keV

  4. Highly multiplexible thermal kinetic inductance detectors for x-ray imaging spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Ulbricht, Gerhard, E-mail: ulbricht@physics.ucsb.edu; Mazin, Benjamin A.; Szypryt, Paul; Walter, Alex B.; Bockstiegel, Clint [Department of Physics, University of California, Santa Barbara, California 93106 (United States); Bumble, Bruce [NASA Jet Propulsion Laboratory, 4800 Oak Grove Drive, Pasadena, California 91125 (United States)

    2015-06-22

    For X-ray imaging spectroscopy, high spatial resolution over a large field of view is often as important as high energy resolution, but current X-ray detectors do not provide both in the same device. Thermal Kinetic Inductance Detectors (TKIDs) are being developed as they offer a feasible way to combine the energy resolution of transition edge sensors with pixel counts approaching CCDs and thus promise significant improvements for many X-ray spectroscopy applications. TKIDs are a variation of Microwave Kinetic Inductance Detectors (MKIDs) and share their multiplexibility: working MKID arrays with 2024 pixels have recently been demonstrated and much bigger arrays are under development. In this work, we present a TKID prototype, which is able to achieve an energy resolution of 75 eV at 5.9 keV, even though its general design still has to be optimized. We further describe TKID fabrication, characterization, multiplexing, and working principle and demonstrate the necessity of a data fitting algorithm in order to extract photon energies. With further design optimizations, we expect to be able to improve our TKID energy resolution to less than 10 eV at 5.9 keV.

  5. Design of dual energy x-ray detector for conveyor belt with steel wire ropes

    Science.gov (United States)

    Dai, Yue; Miao, Changyun; Rong, Feng

    2009-07-01

    A dual energy X-ray detector for conveyor belt with steel wire ropes is researched in the paper. Conveyor belt with steel wire ropes is one of primary transfer equipments in modern production. The traditional test methods like electromagnetic induction principle could not display inner image of steel wire ropes directly. So X-ray detection technology has used to detect the conveyor belt. However the image was not so clear by the interference of the rubber belt. Therefore, the dualenergy X-ray detection technology with subtraction method is developed to numerically remove the rubber belt from radiograph, thus improving the definition of the ropes image. The purpose of this research is to design a dual energy Xray detector that could make the operator easier to found the faulty of the belt. This detection system is composed of Xray source, detector controlled by FPGA chip, PC for running image processing system and so on. With the result of the simulating, this design really improved the capability of the staff to test the conveyor belt.

  6. Energy Calibration of the Pixels of Spectral X-ray Detectors

    CERN Document Server

    Panta, Raj Kumar; Bell, Stephen T; Anderson, Nigel G; Butler, Anthony P; Butler, Philip H

    2015-01-01

    The energy information acquired using spectral X-ray detectors allows noninvasive identification and characterization of chemical components of a material. To achieve this, it is important that the energy response of the detector is calibrated. The established techniques for energy calibration are not practical for routine use in pre-clinical or clinical research environment. This is due to the requirements of using monochromatic radiation sources such as synchrotron, radio-isotopes, and prohibitively long time needed to set up the equipment and make measurements. To address these limitations, we have developed an automated technique for calibrating the energy response of the pixels in a spectral X-ray detector that runs with minimal user intervention. This technique uses the X-ray tube voltage (kVp) as a reference energy, which is stepped through an energy range of interest. This technique locates the energy threshold where a pixel transitions from not-counting (off) to counting (on). Similarly, we have deve...

  7. SENSITIVITY OF STACKED IMAGING DETECTORS TO HARD X-RAY POLARIZATION

    Energy Technology Data Exchange (ETDEWEB)

    Muleri, Fabio; Campana, Riccardo, E-mail: fabio.muleri@iaps.inaf.it [INAF-IAPS, Via del Fosso del Cavaliere 100, I-00133 Roma (Italy)

    2012-06-01

    The development of multi-layer optics which allow to focus photons up to 100 keV and more promises an enormous jump in sensitivity in the hard X-ray energy band. This technology is already planned to be exploited by future missions dedicated to spectroscopy and imaging at energies >10 keV, e.g., Astro-H and NuSTAR. Nevertheless, our understanding of the hard X-ray sky would greatly benefit from carrying out contemporaneous polarimetric measurements, because the study of hard spectral tails and of polarized emission are often two complementary diagnostics of the same non-thermal and acceleration processes. At energies above a few tens of keV, the preferred technique to detect polarization involves the determination of photon directions after a Compton scattering. Many authors have asserted that stacked detectors with imaging capabilities can be exploited for this purpose. If it is possible to discriminate those events which initially interact in the first detector by Compton scattering and are subsequently absorbed by the second layer, then the direction of scattering is singled out from the hit pixels in the two detectors. In this paper, we give the first detailed discussion of the sensitivity of such a generic design to the X-ray polarization. The efficiency and the modulation factor are calculated analytically from the geometry of the instruments and then compared with the performance as derived by means of Geant4 Monte Carlo simulations.

  8. Development of Ta-based Superconducting Tunnel Junction X-ray Detectors for Fluorescence XAS

    International Nuclear Information System (INIS)

    Friedrich, S.; Drury, O.; Hall, J.; Cantor, R.

    2009-01-01

    We are developing superconducting tunnel junction (STJ) soft X-ray detectors for chemical analysis of dilute samples by fluorescence-detected X-ray absorption spectroscopy (XAS). Our 36-pixel Nb-based STJ spectrometer covers a solid angle (Omega)/4π ∼ 10 -3 , offers an energy resolution of ∼10-20 eV FWHM for energies up to ∼1 keV, and can be operated at total count rates of ∼10 6 counts/s. For increased quantum efficiency and cleaner response function, we have now started the development of Ta-based STJ detector arrays. Initial devices modeled after our Nb-based STJs have an energy resolution below 10 eV FWHM for X-ray energies below 1 keV, and pulse rise time discrimination can be used to improve their response function for energies up to several keV. We discuss the performance of the Ta-STJs and outline steps towards the next-generation of large STJ detector arrays with higher sensitivity.

  9. Characterization of the imaging performance of the simultaneously counting and integrating X-ray detector CIX

    Energy Technology Data Exchange (ETDEWEB)

    Fink, Johannes

    2010-01-15

    The CIX detector is a direct converting hybrid pixel detector designed for medical X-ray imaging applications. Its de ning feature is the simultaneous operation of a photon counter as well as an integrator in every pixel cell. This novel approach o ers a dynamic range of more than five orders of magnitude, as well as the ability to directly obtain the average photon energy from the measured data. Several CIX 0.2 ASICs have been successfully connected to CdTe, CdZnTe and Si sensors. These detector modules were tested with respect to the imaging performance of the simultaneously counting and integrating concept under X-ray irradiation. Apart from a characterization of the intrinsic benefits of the CIX concept, the sensor performance was also investigated. Here, the two parallel signal processing concepts offer valuable insights into material related effects like polarization and temporal response. The impact of interpixel coupling effects like charge-sharing, Compton scattering and X-ray fluorescence was evaluated through simulations and measurements. (orig.)

  10. Characterization of the imaging performance of the simultaneously counting and integrating X-ray detector CIX

    International Nuclear Information System (INIS)

    Fink, Johannes

    2010-01-01

    The CIX detector is a direct converting hybrid pixel detector designed for medical X-ray imaging applications. Its de ning feature is the simultaneous operation of a photon counter as well as an integrator in every pixel cell. This novel approach o ers a dynamic range of more than five orders of magnitude, as well as the ability to directly obtain the average photon energy from the measured data. Several CIX 0.2 ASICs have been successfully connected to CdTe, CdZnTe and Si sensors. These detector modules were tested with respect to the imaging performance of the simultaneously counting and integrating concept under X-ray irradiation. Apart from a characterization of the intrinsic benefits of the CIX concept, the sensor performance was also investigated. Here, the two parallel signal processing concepts offer valuable insights into material related effects like polarization and temporal response. The impact of interpixel coupling effects like charge-sharing, Compton scattering and X-ray fluorescence was evaluated through simulations and measurements. (orig.)

  11. Hohlraum Target Alignment from X-ray Detector Images using Starburst Design Patterns

    International Nuclear Information System (INIS)

    Leach, R.R.; Conder, A.; Edwards, O.; Kroll, J.; Kozioziemski, B.; Mapoles, E.; McGuigan, D.; Wilhelmsen, K.

    2010-01-01

    National Ignition Facility (NIF) is a high-energy laser facility comprised of 192 laser beams focused with enough power and precision on a hydrogen-filled spherical, cryogenic target to initiate a fusion reaction. The target container, or hohlraum, must be accurately aligned to an x-ray imaging system to allow careful monitoring of the frozen fuel layer in the target. To achieve alignment, x-ray images are acquired through starburst-shaped windows cut into opposite sides of the hohlraum. When the hohlraum is in alignment, the starburst pattern pairs match nearly exactly and allow a clear view of the ice layer formation on the edge of the target capsule. During the alignment process, x-ray image analysis is applied to determine the direction and magnitude of adjustment required. X-ray detector and source are moved in concert during the alignment process. The automated pointing alignment system described here is both accurate and efficient. In this paper, we describe the control and associated image processing that enables automation of the starburst pointing alignment.

  12. Si(Li) detectors with thin dead layers for low energy x-ray detection

    International Nuclear Information System (INIS)

    Rossington, C.S.; Walton, J.T.; Jaklevic, J.M.

    1990-10-01

    Regions of incomplete charge collection, or ''dead layers'', are compared for Si(Li) detectors fabricated with Au and Pd entrance window electrodes. The dead layers were measured by characterizing the detector spectral response to x-ray energies above and below the Si Kα absorption edge. It was found that Si(Li) detectors with Pd electrodes exhibit consistently thinner effective Si dead layers than those with Au electrodes. Furthermore, it is demonstrated that the minimum thickness required for low resistivity Pd electrodes is thinner than that required for low resistivity Au electrodes, which further reduces the signal attenuation in Pd/Si(Li) detectors. A model, based on Pd compensation of oxygen vacancies in the SiO 2 at the entrance window Si(Li) surface, is proposed to explain the observed differences in detector dead layer thickness. Electrode structures for optimum Si(Li) detector performance at low x-ray energies are discussed. 18 refs., 8 figs., 1 tab

  13. Modelling of the small pixel effect in gallium arsenide X-ray imaging detectors

    CERN Document Server

    Sellin, P J

    1999-01-01

    A Monte Carlo simulation has been carried out to investigate the small pixel effect in highly pixellated X-ray imaging detectors fabricated from semi-insulating gallium arsenide. The presence of highly non-uniform weighting fields in detectors with a small pixel geometry causes the majority of the induced signal to be generated when the moving charges are close to the pixellated contacts. The response of GaAs X-ray imaging detectors is further complicated by the presence of charge trapping, particularly of electrons. In this work detectors are modelled with a pixel pitch of 40 and 150 mu m, and with thicknesses of 300 and 500 mu m. Pulses induced in devices with 40 mu m pixels are due almost totally to the movement of the lightly-trapped holes and can exhibit significantly higher charge collection efficiencies than detectors with large electrodes, in which electron trapping is significant. Details of the charge collection efficiencies as a function of interaction depth in the detector and of the incident phot...

  14. Use of planar HPGe detector as a part of X-ray fluorescent spectrometer for educational purposes

    International Nuclear Information System (INIS)

    Verenchikova, M.S.; Kalinin, V.N.; Mikhajlov, V.A.

    2011-01-01

    This work shows the possibility of use of the nondedicated gamma and X-ray detection head on the basis of planar HPGe detector with a big sensitive area equal to 2000 mm''2 as a part of X-ray fluorescent spectrometer during students' practicum.

  15. Considerations for application of Si(Li) detectors in analyses of sub-keV, ion-induced x rays

    International Nuclear Information System (INIS)

    Musket, R.G.

    1985-01-01

    Spectroscopy of ion-induced x rays is commonly performed using lithium-drifted, silicon detectors, Si(Li), with beryllium windows. Strong absorption of x rays with energies below 1 keV occurs in even the thinnest commercially available beryllium windows and precludes useful analysis of sub-keV x rays. Access to the sub-keV x ray region can be achieved using windowless (WL) and ultra-thin-windowed (UTW) Si(Li) detectors. These detectors have been shown to be useful for spectroscopy of x rays with energies above approximately 200 eV. The properties of such detectors are reviewed with regard to analysis of ion-induced x rays. In particular, considerations of detection efficiency, output linearity, energy resolution, peak shapes, and vacuum requirements are presented. The use of ion excitation for determination of many detector properties serves to demonstrate the usefulness of WL and UTW detectors for the spectroscopy of sub-keV, ion-induced x rays. 23 refs., 4 figs

  16. Computer simulation of the CSPAD, ePix10k, and RayonixMX170HS X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Tina, Adrienne

    2015-08-21

    The invention of free-electron lasers (FELs) has opened a door to an entirely new level of scientific research. The Linac Coherent Light Source (LCLS) at SLAC National Accelerator Laboratory is an X-ray FEL that houses several instruments, each with its own unique X-ray applications. This light source is revolutionary in that while its properties allow for a whole new range of scientific opportunities, it also poses numerous challenges. For example, the intensity of a focused X-ray beam is enough to damage a sample in one mere pulse; however, the pulse speed and extreme brightness of the source together are enough to obtain enough information about that sample, so that no further measurements are necessary. An important device in the radiation detection process, particularly for X-ray imaging, is the detector. The power of the LCLS X-rays has instigated a need for better performing detectors. The research conducted for this project consisted of the study of X-ray detectors to imitate their behaviors in a computer program. The analysis of the Rayonix MX170-HS, CSPAD, and ePix10k in particular helped to understand their properties. This program simulated the interaction of X-ray photons with these detectors to discern the patterns of their responses. A scientist’s selection process of a detector for a specific experiment is simplified from the characterization of the detectors in the program.

  17. VUV and ultrasoft X-ray diode detectors for tokamak plasmas

    International Nuclear Information System (INIS)

    Lee, P.; Snider, R.T.; Gernhardt, J.; Armontrout, C.J.

    1987-11-01

    Ultrasoft X-ray diode (USXRD) arrays have been used on D-IIID and ASDEX to study plasma edge radiation, in the photon energy range from 10 eV to 10 keV. The detectors are extremely useful and versatile due to their simplicity and compactness. Furthermore, absolute quantum efficiencies (QE) of many photocathodes such as vitreous C, Al, Cu CuI, CsI and Au have been measured in recent years. With filter technique, broadband resolution, E/ΔE ≅ 1, is possible. QE comparison of USXRD with semiconductor XRD is also presented to better understand the regions of applicability for each detector. (orig.)

  18. Monte Carlo simulation of the imaging properties of scintillator-coated X-ray pixel detectors

    International Nuclear Information System (INIS)

    Hjelm, M.; Norlin, B.; Nilsson, H.-E.; Froejdh, C.; Badel, X.

    2003-01-01

    The spatial resolution of scintillator-coated X-ray pixel detectors is usually limited by the isotropic light spread in the scintillator. One way to overcome this limitation is to use a pixellated scintillating layer on top of the semiconductor pixel detector. Using advanced etching and filling techniques, arrays of CsI columns have been successfully fabricated and characterized. Each CsI waveguide matches one pixel of the semiconductor detector, limiting the spatial spread of light. Another concept considered in this study is to detect the light emitted from the scintillator by diodes formed in the silicon pore walls. There is so far no knowledge regarding the theoretical limits for these two approaches, which makes the evaluation of the fabrication process difficult. In this work we present numerical calculations of the signal-to-noise ratio (SNR) for detector designs based on scintillator-filled pores in silicon. The calculations are based on separate Monte Carlo (MC) simulations of X-ray absorption and light transport in scintillator waveguides. The resulting data are used in global MC simulations of flood exposures of the detector array, from which the SNR values are obtained. Results are presented for two scintillator materials, namely CsI(Tl) and GADOX

  19. Improvement of an X-ray imaging detector based on a scintillating guides screen

    CERN Document Server

    Badel, X; Linnros, J; Kleimann, P; Froejdh, C; Petersson, C S

    2002-01-01

    An X-ray imaging detector has been developed for dental applications. The principle of this detector is based on application of a silicon charge coupled device covered by a scintillating wave-guide screen. Previous studies of such a detector showed promising results concerning the spatial resolution but low performance in terms of signal to noise ratio (SNR) and sensitivity. Recent results confirm the wave-guiding properties of the matrix and show improvement of the detector in terms of response uniformity, sensitivity and SNR. The present study is focussed on the fabrication of the scintillating screen where the principal idea is to fill a matrix of Si pores with a CsI scintillator. The photoluminescence technique was used to prove the wave-guiding property of the matrix and to inspect the filling uniformity of the pores. The final detector was characterized by X-ray evaluation in terms of spatial resolution, light output and SNR. A sensor with a spatial resolution of 9 LP/mm and a SNR over 50 has been achie...

  20. Mapping the Large Area Straw Detectors of the COMPASS Experiment With X-Rays

    Science.gov (United States)

    Platzer, K.; Dunnweber, W.; Dedek, N.; Faessler, M.; Geyer, R.; Ilgner, C.; Peshekhonov, V.; Wellenstein, H.

    2005-06-01

    In the COMPASS experiment at CERN, large straw drift tube detectors are used for large-angle tracking. To minimize the total areal density, a self supporting structure of thin-walled plastic tubes was chosen and, hence, a loss in mechanical precision was accepted. A complete mapping of the anode wire coordinate grid was required. An X-ray apparatus using a charge-coupled device (CCD) as imaging detector was built to investigate the mechanical properties and to calibrate (offline) the wire positions. Deviations of typically 200-400 /spl mu/m from the nominal positions, defined by equal spacing, are found across the detector area of 8 m/sup 2/. With a calibration method based on high-resolution CCD imaging and pattern recognition algorithms, the absolute wire coordinates are determined with an accuracy better than 30 /spl mu/m across the whole detector area. Temperature effects are clearly seen. Their inhomogenity limits the achievable accuracy to about 50 /spl mu/m under realistic experimental conditions, which is sufficient in view of the intrinsic straw resolution of 200 /spl mu/m for minimum ionizing particles. The offline calibration was checked with particle tracks in the experimental setup, running COMPASS with 160 GeV/c muons. Tracks reconstructed with other detectors that cover a central angular range were used for this comparison. Good agreement is found between these in situ measurements and the X-ray calibration.

  1. Cooled CdZnTe detectors for X-ray astronomy

    CERN Document Server

    Bale, G; Seller, P; Lowe, B

    1999-01-01

    Recent results combining thermoelectrically cooled CdZnTe detectors with a low-noise Pentafet preamplifier are presented. Cooling between -30 deg. C and -40 deg. C reduces the leakage current of the detectors and allows the use of a pulsed reset preamplifier and long shaping times, significantly improving the energy resolution. Mn K subalpha X-rays at 5.9 keV have been observed with a resolution of less than 280 eV FWHM and a peak to background of more than 200:1. The Fano factor of the material has been estimated at 0.11+-0.012 at -40 deg. C. The detector requirement for X-ray astronomy will be a photon-counting imaging spectrometer. A 16x16 element, bump bonded pixel detector is described and results from a prototype silicon array presented. The detector is constructed with ASIC amplifiers with a system noise of <25 electrons rms and should give an energy resolution comparable to the Pentafet results presented here.

  2. Serial data acquisition for the X-ray plasma diagnostics with selected GEM detector structures

    International Nuclear Information System (INIS)

    Czarski, T.; Chernyshova, M.; Pozniak, K.T.; Kasprowicz, G.; Zabolotny, W.; Kolasinski, P.; Krawczyk, R.; Wojenski, A.; Zienkiewicz, P.

    2015-01-01

    The measurement system based on GEM—Gas Electron Multiplier detector is developed for X-ray diagnostics of magnetic confinement tokamak plasmas. The paper is focused on the measurement subject and describes the fundamental data processing to obtain reliable characteristics (histograms) useful for physicists. The required data processing have two steps: 1—processing in the time domain, i.e. events selections for bunches of coinciding clusters, 2—processing in the planar space domain, i.e. cluster identification for the given detector structure. So, it is the software part of the project between the electronic hardware and physics applications. The whole project is original and it was developed by the paper authors. The previous version based on 1-D GEM detector was applied for the high-resolution X-ray crystal spectrometer KX1 in the JET tokamak. The current version considers 2-D detector structures for the new data acquisition system. The fast and accurate mode of data acquisition implemented in the hardware in real time can be applied for the dynamic plasma diagnostics. Several detector structures with single-pixel sensors and multi-pixel (directional) sensors are considered for two-dimensional X-ray imaging. Final data processing is presented by histograms for selected range of position, time interval and cluster charge values. Exemplary radiation source properties are measured by the basic cumulative characteristics: the cluster position distribution and cluster charge value distribution corresponding to the energy spectra. A shorter version of this contribution is due to be published in PoS at: 1 st EPS conference on Plasma Diagnostics

  3. Serial data acquisition for the X-ray plasma diagnostics with selected GEM detector structures

    Science.gov (United States)

    Czarski, T.; Chernyshova, M.; Pozniak, K. T.; Kasprowicz, G.; Zabolotny, W.; Kolasinski, P.; Krawczyk, R.; Wojenski, A.; Zienkiewicz, P.

    2015-10-01

    The measurement system based on GEM—Gas Electron Multiplier detector is developed for X-ray diagnostics of magnetic confinement tokamak plasmas. The paper is focused on the measurement subject and describes the fundamental data processing to obtain reliable characteristics (histograms) useful for physicists. The required data processing have two steps: 1—processing in the time domain, i.e. events selections for bunches of coinciding clusters, 2—processing in the planar space domain, i.e. cluster identification for the given detector structure. So, it is the software part of the project between the electronic hardware and physics applications. The whole project is original and it was developed by the paper authors. The previous version based on 1-D GEM detector was applied for the high-resolution X-ray crystal spectrometer KX1 in the JET tokamak. The current version considers 2-D detector structures for the new data acquisition system. The fast and accurate mode of data acquisition implemented in the hardware in real time can be applied for the dynamic plasma diagnostics. Several detector structures with single-pixel sensors and multi-pixel (directional) sensors are considered for two-dimensional X-ray imaging. Final data processing is presented by histograms for selected range of position, time interval and cluster charge values. Exemplary radiation source properties are measured by the basic cumulative characteristics: the cluster position distribution and cluster charge value distribution corresponding to the energy spectra. A shorter version of this contribution is due to be published in PoS at: 1st EPS conference on Plasma Diagnostics

  4. On-board event processing algorithms for a CCD-based space borne X-ray spectrometer

    International Nuclear Information System (INIS)

    Chun, H.J.; Bowles, J.A.; Branduardi-Raymont, G.; Gowen, R.A.

    1996-01-01

    This paper describes two alternative algorithms which are applied to reduce the telemetry requirements for a Charge Coupled Device (CCD) based, space-borne, X-ray spectrometer by on-board reconstruction of the X-ray events split over two or more adjacent pixels. The algorithms have been developed for the Reflection Grating Spectrometer (RGS) on the X-ray multi-mirror (XMM) mission, the second cornerstone project in the European Space Agency's Horizon 2000 programme. The overall instrument and some criteria which provide the background of the development of the algorithms, implemented in Tartan ADA on an MA31750 microprocessor, are described. The on-board processing constraints and requirements are discussed, and the performances of the algorithms are compared. Test results are presented which show that the recursive implementation is faster and has a smaller executable file although it uses more memory because of its stack requirements. (orig.)

  5. Fabrication of a Tantalum-Based Josephson Junction for an X-Ray Detector

    Science.gov (United States)

    Morohashi, Shin'ichi; Gotoh, Kohtaroh; Yokoyama, Naoki

    2000-06-01

    We have fabricated a tantalum-based Josephson junction for an X-ray detector. The tantalum layer was selected for the junction electrode because of its long quasiparticle lifetime, large X-ray absorption efficiency and stability against thermal cycling. We have developed a buffer layer to fabricate the tantalum layer with a body-centered cubic structure. Based on careful consideration of their superconductivity, we have selected a niobium thin layer as the buffer layer for fabricating the tantalum base electrode, and a tungsten thin layer for the tantalum counter electrode. Fabricated Nb/AlOx-Al/Ta/Nb and Nb/Ta/W/AlOx-Al/Ta/Nb Josephson junctions exhibited current-voltage characteristics with a low subgap leakage current.

  6. CdTe in photoconductive applications. Fast detector for metrology and X-ray imaging

    International Nuclear Information System (INIS)

    Cuzin, M.

    1991-01-01

    Operating as a photoconductor, the sensitivity and the impulse response of semi-insulating materials greatly depend on the excitation duration compared to electron and hole lifetimes. The requirement of ohmic contact is shortly discussed. Before developing picosecond measurements with integrated autocorrelation system, this paper explains high energy industrial tomographic application with large CdTe detectors (25x15x0.9 mm 3 ). The excitation is typically μs range. X-ray flash radiography, with 10 ns burst, is in an intermediate time domain where excitation is similar to electron life-time. In laser fusion experiment excitation is in the range of 50 ps and we develop photoconductive devices able to study very high speed X-ray emission time behaviour. Thin polycristalline MOCVD CdTe films with picosecond response are suitable to perform optical correlation measurements of single shot pulses with a very large bandwidth (- 50 GHz)

  7. Observation of superheavy primary cosmic ray nuclei with solid state track detectors and x-ray films

    International Nuclear Information System (INIS)

    Doke, Tadayoshi; Hayashi, Takayoshi; Ito, Kensai; Yanagimachi, Tomoki; Kobayashi, Shigeru.

    1977-01-01

    The measurements of energy spectra and the nuclear charge distribution of superheavy nuclei heavier than iron in primary cosmic ray can provide information on the origin, propagation and life time of the cosmic ray. Since incident particles are in the region of relativistic velocity (the low energy cosmic ray below the cutoff energy is forbidden from entering), the charges of cosmic ray nuclei can be determined without knowing the energy of particles. The balloon-borne solid state track detector and plastic and X-ray films were employed for the detection of superheavy cosmic ray, and the five events were detected with the cellulose nitrate film. The flux of superheavy nuclei is predicted from the present analysis. (Yoshimori, M.)

  8. Characterizing X-ray detectors for prototype digital breast tomosynthesis systems

    International Nuclear Information System (INIS)

    Kim, Y.-S.; Park, H.-S.; Park, S.-J.; Choi, S.; Lee, H.; Kim, H.-J.; Lee, D.; Choi, Y.-W.

    2016-01-01

    The digital breast tomosynthesis (DBT) system is a newly developed 3-D imaging technique that overcomes the tissue superposition problems of conventional mammography. Therefore, it produces fewer false positives. In DBT system, several parameters are involved in image acquisition, including geometric components. A series of projections should be acquired at low exposure. This makes the system strongly dependent on the detector's characteristic performance. This study compares two types of x-ray detectors developed by the Korea Electrotechnology Research Institute (KERI). The first prototype DBT system has a CsI (Tl) scintillator/CMOS based flat panel digital detector (2923 MAM, Dexela Ltd.), with a pixel size of 0.0748 mm. The second uses a-Se based direct conversion full field detector (AXS 2430, analogic) with a pixel size of 0.085 mm. The geometry of both systems is same, with a focal spot 665.8 mm from the detector, and a center of rotation 33 mm above the detector surface. The systems were compared with regard to modulation transfer function (MTF), normalized noise power spectrum (NNPS), detective quantum efficiency (DQE) and a new metric, the relative object detectability (ROD). The ROD quantifies the relative performance of each detector at detecting specified objects. The system response function demonstrated excellent linearity (R 2 >0.99). The CMOS-based detector had a high sensitivity, while the Anrad detector had a large dynamic range. The higher MTF and noise power spectrum (NPS) values were measured using an Anrad detector. The maximum DQE value of the Dexela detector was higher than that of the Anrad detector with a low exposure level, considering one projection exposure for tomosynthesis. Overall, the Dexela detector performed better than did the Anrad detector with regard to the simulated Al wires, spheres, test objects of ROD with low exposure level. In this study, we compared the newly developed prototype DBT system with two different types

  9. Neutron Radiation Shielding For The NIF Streaked X-Ray Detector (SXD) Diagnostic

    International Nuclear Information System (INIS)

    Song, P; Holder, J; Young, B; Kalantar, D; Eder, D; Kimbrough, J

    2006-01-01

    The National Ignition Facility (NIF) at Lawrence Livermore National Laboratory (LLNL) is preparing for the National Ignition Campaign (NIC) scheduled in 2010. The NIC is comprised of several ''tuning'' physics subcampaigns leading up to a demonstration of Inertial Confinement Fusion (ICF) ignition. In some of these experiments, time-resolved x-ray imaging of the imploding capsule may be required to measure capsule trajectory (shock timing) or x-ray ''bang-time''. A capsule fueled with pure tritium (T) instead of a deutriun-tritium (DT) mixture is thought to offer useful physics surrogacy, with reduced yields of up to 5e14 neutrons. These measurements will require the use of the NIF streak x-ray detector (SXD). The resulting prompt neutron fluence at the planned SXD location (∼1.7 m from the target) would be ∼1.4e9/cm 2 . Previous measurements suggest the onset of significant background at a neutron fluence of ∼ 1e8/cm 2 . The radiation damage and operational upsets which starts at ∼1e8 rad-Si/sec must be factored into an integrated experimental campaign plan. Monte Carlo analyses were performed to predict the neutron and gamma/x-ray fluences and radiation doses for the proposed diagnostic configuration. A possible shielding configuration is proposed to mitigate radiation effects. The primary component of this shielding is an 80 cm thickness of Polyethylene (PE) between target chamber center (TCC) and the SXD diagnostic. Additionally, 6-8 cm of PE around the detector provide from the large number of neutrons that scatter off the inside of the target chamber. This proposed shielding configuration reduces the high-energy neutron fluence at the SXD by approximately a factor ∼50

  10. Neutron Radiation Shielding For The NIF Streaked X-Ray Detector (SXD) Diagnostic

    Energy Technology Data Exchange (ETDEWEB)

    Song, P; Holder, J; Young, B; Kalantar, D; Eder, D; Kimbrough, J

    2006-11-02

    The National Ignition Facility (NIF) at Lawrence Livermore National Laboratory (LLNL) is preparing for the National Ignition Campaign (NIC) scheduled in 2010. The NIC is comprised of several ''tuning'' physics subcampaigns leading up to a demonstration of Inertial Confinement Fusion (ICF) ignition. In some of these experiments, time-resolved x-ray imaging of the imploding capsule may be required to measure capsule trajectory (shock timing) or x-ray ''bang-time''. A capsule fueled with pure tritium (T) instead of a deutriun-tritium (DT) mixture is thought to offer useful physics surrogacy, with reduced yields of up to 5e14 neutrons. These measurements will require the use of the NIF streak x-ray detector (SXD). The resulting prompt neutron fluence at the planned SXD location ({approx}1.7 m from the target) would be {approx}1.4e9/cm{sup 2}. Previous measurements suggest the onset of significant background at a neutron fluence of {approx} 1e8/cm{sup 2}. The radiation damage and operational upsets which starts at {approx}1e8 rad-Si/sec must be factored into an integrated experimental campaign plan. Monte Carlo analyses were performed to predict the neutron and gamma/x-ray fluences and radiation doses for the proposed diagnostic configuration. A possible shielding configuration is proposed to mitigate radiation effects. The primary component of this shielding is an 80 cm thickness of Polyethylene (PE) between target chamber center (TCC) and the SXD diagnostic. Additionally, 6-8 cm of PE around the detector provide from the large number of neutrons that scatter off the inside of the target chamber. This proposed shielding configuration reduces the high-energy neutron fluence at the SXD by approximately a factor {approx}50.

  11. Evaluation of a photon counting Medipix3RX CZT spectral x-ray detector

    Science.gov (United States)

    Jorgensen, Steven M.; Vercnocke, Andrew J.; Rundle, David S.; Butler, Philip H.; McCollough, Cynthia H.; Ritman, Erik L.

    2016-10-01

    We assessed the performance of a cadmium zinc telluride (CZT)-based Medipix3RX x-ray detector as a candidate for micro-computed tomography (micro-CT) imaging. This technology was developed at CERN for the Large Hadron Collider. It features an array of 128 by 128, 110 micrometer square pixels, each with eight simultaneous threshold counters, five of which utilize real-time charge summing, significantly reducing the charge sharing between contiguous pixels. Pixel response curves were created by imaging a range of x-ray intensities by varying x-ray tube current and by varying the exposure time with fixed x-ray current. Photon energy-related assessments were made by flooding the detector with the tin foil filtered emission of an I-125 radioisotope brachytherapy seed and sweeping the energy threshold of each of the four charge-summed counters of each pixel in 1 keV steps. Long term stability assessments were made by repeating exposures over the course of one hour. The high properly-functioning pixel yield (99%), long term stability (linear regression of whole-chip response over one hour of acquisitions: y = -0.0038x + 2284; standard deviation: 3.7 counts) and energy resolution (2.5 keV FWHM (single pixel), 3.7 keV FWHM across the full image) make this device suitable for spectral micro-CT. The charge summing performance effectively reduced the measurement corruption caused by charge sharing which, when unaccounted for, shifts the photon energy assignment to lower energies, degrading both count and energy accuracy. Effective charge summing greatly improves the potential for calibrated, energy-specific material decomposition and K edge difference imaging approaches.

  12. Silicon drift detectors with on-chip electronics for x-ray spectroscopy.

    Science.gov (United States)

    Fiorini, C; Longoni, A; Hartmann, R; Lechner, P; Strüder, L

    1997-01-01

    The silicon drift detector (SDD) is a semiconductor device based on high resistivity silicon fully depleted through junctions implanted on both sides of the semiconductor wafer. The electrons generated by the ionizing radiation are driven by means of a suitable electric field from the point of interaction toward a collecting anode of small capacitance, independent of the active area of the detector. A suitably designed front-end JFET has been directly integrated on the detector chip close to the anode region, in order to obtain a nearly ideal capacitive matching between detector and transistor and to minimize the stray capacitances of the connections. This feature allows it to reach high energy resolution also at high count rates and near room temperature. The present work describes the structure and the performance of SDDs specially designed for high resolution spectroscopy with soft x rays at high detection rate. Experimental results of SDDs used in spectroscopy applications are also reported.

  13. Possible use of CdTe detectors in kVp monitoring of diagnostic X-ray tubes

    International Nuclear Information System (INIS)

    Krmar, M.; Bucalovic, N.; Baucal, M.; Jovancevic, N.

    2010-01-01

    It has been suggested that kVp of diagnostic X-ray devices (or maximal energy of X-ray photon spectra) should be monitored routinely; however a standardized non-invasive technique has yet to be developed and proposed. It is well known that the integral number of Compton scattered photons and the intensities of fluorescent X-ray lines registered after irradiation of some material by an X-ray beam are a function of the maximal beam energy. CdTe detectors have sufficient energy resolution to distinguish individual X-ray fluorescence lines and high efficiency for the photon energies in the diagnostic region. Our initial measurements have demonstrated that the different ratios of the integral number of Compton scattered photons and intensities of K and L fluorescent lines detected by CdTe detector are sensitive function of maximal photon energy and could be successfully applied for kVp monitoring.

  14. Uncooled Radiation Hard Large Area SiC X-ray and EUV Detectors and 2D Arrays, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — This project seeks to design, fabricate, characterize and commercialize large area, uncooled and radiative hard 4H-SiC EUV ? soft X-ray detectors capable of ultra...

  15. Analysis and operation of DePFET X-ray imaging detectors

    International Nuclear Information System (INIS)

    Lauf, Thomas

    2011-01-01

    The latest active pixel sensor for X-ray imaging spectroscopy developed at the Max-Planck-Halbleiterlabor (HLL) is the Depleted P-channel Field Effect Transistor (DePFET). This detector type unites detector and first stage amplification and has excellent energy resolution, low noise readout at high speed and low power consumption. This is combined with the possibility of random accessibility of pixels and on-demand readout. In addition it possesses all advantages of a sidewards depleted device, i.e. 100% fill factor and very good quantum efficiency. In the course of the development of DePFET detectors the need of a data analysis software for DePFET devices became apparent. A new tool was developed within the scope of this thesis, which should enable scientists to analyze DePFET data, but also be flexible enough so it can be adapted to new device variants and analysis challenges. A modular concept was thus implemented: a base program running an analysis by individual steps encapsulating algorithms, which can be interchanged. The result is a flexible, adaptable, and expandable analysis software. The software was used to investigate and qualify different structural variants of DePFET detectors. Algorithms to examine detector effects and methods to correct them were developed and integrated into the software. This way, a standard analysis suite for DePFET data was built up which is used at the HLL. Beside the planned use as detector for the wide field imager in the space X-ray observatory IXO, DePFET matrices will be used as focal plane array on the Mercury Imaging X-ray Spectrometer on board the Mercury probe BepiColombo which is scheduled for launch in 2014. The developed analysis software was used in the detector development for this mission to qualify test structures, analyze detector effects and study experimental results. In the course of this development, detector prototypes were studied in respect of linearity, charge collection and detection efficiency in an

  16. Analysis and operation of DePFET X-ray imaging detectors

    Energy Technology Data Exchange (ETDEWEB)

    Lauf, Thomas

    2011-04-28

    The latest active pixel sensor for X-ray imaging spectroscopy developed at the Max-Planck-Halbleiterlabor (HLL) is the Depleted P-channel Field Effect Transistor (DePFET). This detector type unites detector and first stage amplification and has excellent energy resolution, low noise readout at high speed and low power consumption. This is combined with the possibility of random accessibility of pixels and on-demand readout. In addition it possesses all advantages of a sidewards depleted device, i.e. 100% fill factor and very good quantum efficiency. In the course of the development of DePFET detectors the need of a data analysis software for DePFET devices became apparent. A new tool was developed within the scope of this thesis, which should enable scientists to analyze DePFET data, but also be flexible enough so it can be adapted to new device variants and analysis challenges. A modular concept was thus implemented: a base program running an analysis by individual steps encapsulating algorithms, which can be interchanged. The result is a flexible, adaptable, and expandable analysis software. The software was used to investigate and qualify different structural variants of DePFET detectors. Algorithms to examine detector effects and methods to correct them were developed and integrated into the software. This way, a standard analysis suite for DePFET data was built up which is used at the HLL. Beside the planned use as detector for the wide field imager in the space X-ray observatory IXO, DePFET matrices will be used as focal plane array on the Mercury Imaging X-ray Spectrometer on board the Mercury probe BepiColombo which is scheduled for launch in 2014. The developed analysis software was used in the detector development for this mission to qualify test structures, analyze detector effects and study experimental results. In the course of this development, detector prototypes were studied in respect of linearity, charge collection and detection efficiency in an

  17. Application of GaAs and CdTe photoconductor detectors to x-ray flash radiography

    International Nuclear Information System (INIS)

    Mathy, F.; Cuzin, M.; Gagelin, J.J.; Mermet, R.; Piaget, B.; Rustique, J.; Verger, L.

    1991-01-01

    Semi-insulating GaAs and CdTe:Cl photoconductor probes were qualified on high energy X ray single shot flash generators. The estimated minimum detected dose per flash corresponding to a 230 mrad direct beam attenuated by 200 mm lead was 20 μrad. The dynamic range was about 4 decades in amplitude or charges, with a good linearity. Such detectors, by locating the origin of the parasitic scattered beam, could be used to eliminate this parasitic beam in X ray flash radiography in detonics experiments. Imaging possibilities are mentioned, as well as X ray generator monitoring with such detectors or with neutron preirradiated photoconductors

  18. Application of GaAs and CdTe photoconductor detectors to X-ray flash radiography

    International Nuclear Information System (INIS)

    Mathy, F.; Cuzin, M.; Gagelin, J.J.; Mermet, R.; Piaget, B.; Rustique, J.; Verger, L.; Hauducoeur, A.; Nicolas, P.; Le Dain, L.; Hyvernage, M.

    1992-01-01

    Some insulating GaAs and CdTe:Cl photoconductor probes were qualified on high energy X-ray single-shot flash generators. The estimated minimum detected dose per flash corresponding to a 230 mrad direct beam attenuated by 200 mm lead was 20 μrad. The dynamic range was about 4 decades in amplitude or charge, with a good linearity. Such detectors, by locating the origin of the parasitic scattered beam, could be used to eliminate this parasitic beam in X-ray flash radiography in detonics experiments. Imaging possibilities are mentioned, as well as X-ray generator monitoring with such detectors or with neutron preirradiated photoconductors. (orig.)

  19. Application of GaAs and CdTe photoconductor detectors to X-ray flash radiography

    Energy Technology Data Exchange (ETDEWEB)

    Mathy, F.; Cuzin, M.; Gagelin, J.J.; Mermet, R.; Piaget, B.; Rustique, J.; Verger, L. (CEA, Direction des Technologies Avancees, Lab. d' Electronique, de Technologie et d' Instrumentation, DSYS, 38 - Grenoble (France)); Hauducoeur, A.; Nicolas, P.; Le Dain, L.; Hyvernage, M. (CEA, Direction des Applications Militaires, 77 - Courtry (France))

    1992-11-15

    Some insulating GaAs and CdTe:Cl photoconductor probes were qualified on high energy X-ray single-shot flash generators. The estimated minimum detected dose per flash corresponding to a 230 mrad direct beam attenuated by 200 mm lead was 20 [mu]rad. The dynamic range was about 4 decades in amplitude or charge, with a good linearity. Such detectors, by locating the origin of the parasitic scattered beam, could be used to eliminate this parasitic beam in X-ray flash radiography in detonics experiments. Imaging possibilities are mentioned, as well as X-ray generator monitoring with such detectors or with neutron preirradiated photoconductors. (orig.).

  20. Application of GaAs and CdTe photoconductor detectors to x-ray flash radiography

    Energy Technology Data Exchange (ETDEWEB)

    Mathy, F.; Cuzin, M.; Gagelin, J.J.; Mermet, R.; Piaget, B.; Rustique, J.; Verger, L. [CEA Centre d`Etudes de Grenoble, 38 (FR). Direction des Technologies Avancees; Hauducoeur, A.; Nicolas, P.; Le Dain, L.; Hyvernage, M. [CEA Centre d`Etudes de Vaujours, 77 - Courtry (FR)

    1991-12-31

    Semi-insulating GaAs and CdTe:Cl photoconductor probes were qualified on high energy X ray single shot flash generators. The estimated minimum detected dose per flash corresponding to a 230 mrad direct beam attenuated by 200 mm lead was 20 {mu}rad. The dynamic range was about 4 decades in amplitude or charges, with a good linearity. Such detectors, by locating the origin of the parasitic scattered beam, could be used to eliminate this parasitic beam in X ray flash radiography in detonics experiments. Imaging possibilities are mentioned, as well as X ray generator monitoring with such detectors or with neutron preirradiated photoconductors.

  1. The application of photoconductive detectors to the measurement of x-ray production in laser produced plasmas

    International Nuclear Information System (INIS)

    Kania, D.R.; Bell, P.; Trebes, J.

    1987-08-01

    Photoconductive detectors (PCDs) offer an attractive alternative for the measurement of pulsed x-rays from laser produced plasmas. These devices are fast (FWHM ∼100 ps), sensitive and simple to use. We have used InP, GaAs, and Type IIa diamond as PCDs to measure x-rays emission from 100 eV to 100 keV. Specifically, we have used these detectors to measure total radiation yields, corona temperatures, and hot electron generated x-rays from laser produced plasmas. 5 refs., 4 figs

  2. Evaluation of radiation tolerance of FETs used for Astro-E2 hard X-ray detector (HXD-II)

    International Nuclear Information System (INIS)

    Itoh, Takeshi; Niko, Hisako; Kokubun, Motohide; Makishima, Kazuo; Kawaharada, Madoka; Takahashi, Isao; Miyasaka, Hiromasa

    2005-01-01

    We evaluated the radiation tolerance of three types of metal-can MOS Field Effect Transistors (FETs). They are candidates for flight electronics of the Hard X-ray Detector (HXD-II) experiment which is onboard the cosmic X-ray satellite Astro-E2 scheduled for launch in 2005. We irradiated FETs with a Co60γ-ray source under several different experimental conditions, and measured changes in their I-V characteristic curves. After a 10krad irradiation during which the gate voltage is set at 0V, all types showed a decrease in the switching voltage by ∼0.2-0.4V. In addition, the gate conductance increased under some irradiation conditions. These experimental results may be explained in terms of trapped charges and boundary levels in the oxide layer beneath the gate electrode. We have confirmed that at least two types of FETs can be used in our satellite-borne experiment, one as relay-driving FETs and the other in TTL-ECL conversion circuits

  3. Modeling the frequency-dependent detective quantum efficiency of photon-counting x-ray detectors.

    Science.gov (United States)

    Stierstorfer, Karl

    2018-01-01

    To find a simple model for the frequency-dependent detective quantum efficiency (DQE) of photon-counting detectors in the low flux limit. Formula for the spatial cross-talk, the noise power spectrum and the DQE of a photon-counting detector working at a given threshold are derived. Parameters are probabilities for types of events like single counts in the central pixel, double counts in the central pixel and a neighboring pixel or single count in a neighboring pixel only. These probabilities can be derived in a simple model by extensive use of Monte Carlo techniques: The Monte Carlo x-ray propagation program MOCASSIM is used to simulate the energy deposition from the x-rays in the detector material. A simple charge cloud model using Gaussian clouds of fixed width is used for the propagation of the electric charge generated by the primary interactions. Both stages are combined in a Monte Carlo simulation randomizing the location of impact which finally produces the required probabilities. The parameters of the charge cloud model are fitted to the spectral response to a polychromatic spectrum measured with our prototype detector. Based on the Monte Carlo model, the DQE of photon-counting detectors as a function of spatial frequency is calculated for various pixel sizes, photon energies, and thresholds. The frequency-dependent DQE of a photon-counting detector in the low flux limit can be described with an equation containing only a small set of probabilities as input. Estimates for the probabilities can be derived from a simple model of the detector physics. © 2017 American Association of Physicists in Medicine.

  4. A diamond detector for inertial confinement fusion X-ray bang-time measurements at the National Ignition Facility

    Energy Technology Data Exchange (ETDEWEB)

    MacPhee, A G; Brown, C; Burns, S; Celeste, J; Glenzer, S H; Hey, D; Jones, O S; Landen, O; Mackinnon, A J; Meezan, N; Parker, J; Edgell, D; Glebov, V Y; Kilkenny, J; Kimbrough, J

    2010-11-09

    An instrument has been developed to measure X-ray bang-time for inertial confinement fusion capsules; the time interval between the start of the laser pulse and peak X-ray emission from the fuel core. The instrument comprises chemical vapor deposited polycrystalline diamond photoconductive X-ray detectors with highly ordered pyrolytic graphite X-ray monochromator crystals at the input. Capsule bang-time can be measured in the presence of relatively high thermal and hard X-ray background components due to the selective band pass of the crystals combined with direct and indirect X-ray shielding of the detector elements. A five channel system is being commissioned at the National Ignition Facility at Lawrence Livermore National Laboratory for implosion optimization measurements as part of the National Ignition Campaign. Characteristics of the instrument have been measured demonstrating that X-ray bang-time can be measured with {+-} 30ps precision, characterizing the soft X-ray drive to +/- 1eV or 1.5%.

  5. Aerogel Cherenkov detector for characterizing the intense flash x-ray source, Cygnus, spectrum

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Y., E-mail: yhkim@lanl.gov; Herrmann, H. W.; McEvoy, A. M.; Young, C. S.; Hamilton, C. [Los Alamos National Laboratory, Los Alamos, New Mexico 87545 (United States); Schwellenbach, D. D.; Malone, R. M.; Kaufman, M. I.; Smith, A. S. [National Security Technologies, LLC, Los Alamos, New Mexico 87544 (United States)

    2016-11-15

    An aerogel Cherenkov detector is proposed to measure the X-ray energy spectrum from the Cygnus—intense flash X-ray source operated at the Nevada National Security Site. An array of aerogels set at a variety of thresholds between 1 and 3 MeV will be adequate to map out the bremsstrahlung X-ray production of the Cygnus, where the maximum energy of the spectrum is normally around 2.5 MeV. In addition to the Cherenkov radiation from aerogels, one possible competing light-production mechanism is optical transition radiation (OTR), which may be significant in aerogels due to the large number of transitions from SiO{sub 2} clusters to vacuum voids. To examine whether OTR is a problem, four aerogel samples were tested using a mono-energetic electron beam (varied in the range of 1–3 MeV) at NSTec Los Alamos Operations. It was demonstrated that aerogels can be used as a Cherenkov medium, where the rate of the light production is about two orders magnitude higher when the electron beam energy is above threshold.

  6. Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensors

    International Nuclear Information System (INIS)

    Koenig, Thomas; Schulze, Julia; Zuber, Marcus; Rink, Kristian; Oelfke, Uwe; Butzer, Jochen; Hamann, Elias; Cecilia, Angelica; Zwerger, Andreas; Fauler, Alex; Fiederle, Michael

    2012-01-01

    Spectroscopic x-ray imaging by means of photon counting detectors has received growing interest during the past years. Critical to the image quality of such devices is their pixel pitch and the sensor material employed. This paper describes the imaging properties of Medipix2 MXR multi-chip assemblies bump bonded to 1 mm thick CdTe sensors. Two systems were investigated with pixel pitches of 110 and 165 μm, which are in the order of the mean free path lengths of the characteristic x-rays produced in their sensors. Peak widths were found to be almost constant across the energy range of 10 to 60 keV, with values of 2.3 and 2.2 keV (FWHM) for the two pixel pitches. The average number of pixels responding to a single incoming photon are about 1.85 and 1.45 at 60 keV, amounting to detective quantum efficiencies of 0.77 and 0.84 at a spatial frequency of zero. Energy selective CT acquisitions are presented, and the two pixel pitches' abilities to discriminate between iodine and gadolinium contrast agents are examined. It is shown that the choice of the pixel pitch translates into a minimum contrast agent concentration for which material discrimination is still possible. We finally investigate saturation effects at high x-ray fluxes and conclude with the finding that higher maximum count rates come at the cost of a reduced energy resolution. (paper)

  7. Parametric curve evaluation of a phototransistor used as detector in stereotactic radiosurgery X-ray beam

    International Nuclear Information System (INIS)

    Lima, Daniela Pontes A.; Santos, Luiz Antonio P.; Santos, Walter M.; Silva Junior, Eronides F. da

    2005-01-01

    Phototransistors have been widely used as detectors for low energy X-rays. However, when they are used in high energy X-rays fields like those generated from linear accelerators (linac), there is a certain loss of sensibility to the ionizing radiation. This damage is cumulative and irreversible. Thus, a correction factor must be applied to its response, which is proportional to the integrated dose. However, it is possible to estimate the correction factor by using the V x I parametric curve of the device. The aim of this work was to develop studies to evaluate and correlate the parametric response curve of a phototransistor with its loss of sensibility after irradiation. An Agilent 4155C semiconductor parameter analyzer was used to trace the parametric curve. X-rays were generated by a 14 MV Primus-Siemens linear accelerator. The results demonstrated that there is a correlation between the integrated dose applied to the phototransistor and the parametric response of the device. Studies are under way to determine how such behavior can provide information for the dosimetric planning in stereotactic radiosurgery. (author)

  8. Collecting data in the home laboratory: evolution of X-ray sources, detectors and working practices

    Energy Technology Data Exchange (ETDEWEB)

    Skarzynski, Tadeusz, E-mail: tadeusz.skarzynski@agilent.com [Agilent Technologies, 10 Mead Road, Yarnton, Oxfordshire (United Kingdom)

    2013-07-01

    Recent developments in X-ray crystallographic hardware related to structural biology research are presented and discussed. While the majority of macromolecular X-ray data are currently collected using highly efficient beamlines at an ever-increasing number of synchrotrons, there is still a need for high-performance reliable systems for in-house experiments. In addition to crystal screening and optimization of data-collection parameters before a synchrotron trip, the home system allows the collection of data as soon as the crystals are produced to obtain the solution of novel structures, especially by the molecular-replacement method, and is invaluable in achieving the quick turnover that is often required for ligand-binding studies in the pharmaceutical industry. There has been a continuous evolution of X-ray sources, detectors and software developed for in-house use in recent years and a diverse range of tools for structural biology laboratories are available. An overview of the main directions of these developments and examples of specific solutions available to the macromolecular crystallography community are presented in this paper, showing that data collection ‘at home’ is still an attractive proposition complementing the use of synchrotron beamlines.

  9. Characteristics and quality test of X-ray with CZT detector

    International Nuclear Information System (INIS)

    Rianto, Sugeng

    2000-01-01

    The study examines the use of direct measurements of x-ray spectra for testing the quality of x-ray beam by using a peltier-cooled CZT detector under different conditions. The typical calibration of the spectrometry system shows that the energy resolution of the system is 1,2 keV at 122 keV of 57 Co. The utilization CZT based spectrometer for assessing the quality of x-ray machine on its spectra show that the CZT could accurately measure the spectra at various kVp, m As and filtration, except at the kVp greater than 140 and m As higher than 2 without added filtration. A comparison of CZT with the field instruments showed that there is a reasonable agreement between the Keithley and CZT at lower energies regardless of filtration however at high energies there is a large difference.In contrast the discrepancy between the CZT and Nero increased at lower photon energies particularly for high filtration

  10. Development of an ultra-fast X-ray camera using hybrid pixel detectors

    International Nuclear Information System (INIS)

    Dawiec, A.

    2011-05-01

    The aim of the project whose work described in this thesis is part, was to design a high-speed X-ray camera using hybrid pixels applied to biomedical imaging and for material science. As a matter of fact the hybrid pixel technology meets the requirements of these two research fields, particularly by providing energy selection and low dose imaging capabilities. In this thesis, high frame rate X-ray imaging based on the XPAD3-S photons counting chip is presented. Within a collaboration between CPPM, ESRF and SOLEIL, three XPAD3 cameras were built. Two of them are being operated at the beamline of the ESRF and SOLEIL synchrotron facilities and the third one is embedded in the PIXSCAN II irradiation setup of CPPM. The XPAD3 camera is a large surface X-ray detector composed of eight detection modules of seven XPAD3-S chips each with a high-speed data acquisition system. The readout architecture of the camera is based on the PCI Express interface and on programmable FPGA chips. The camera achieves a readout speed of 240 images/s, with maximum number of images limited by the RAM memory of the acquisition PC. The performance of the device was characterized by carrying out several high speed imaging experiments using the PIXSCAN II irradiation setup described in the last chapter of this thesis. (author)

  11. Sputtered carbon as a corrosion barrier for x-ray detector windows

    Energy Technology Data Exchange (ETDEWEB)

    Rowley, Joseph; Pei, Lei; Davis, Robert C., E-mail: davis@byu.edu; Vanfleet, Richard R. [Department of Physics and Astronomy, Brigham Young University, Provo, Utah 84602 (United States); Liddiard, Steven; Harker, Mallorie; Abbott, Jonathan [Moxtek, Inc., 452 W 1260 N, Orem, Utah 84057 (United States)

    2016-09-15

    Sputtered amorphous carbon thin films were explored as corrosion resistant coatings on aluminum thin films to be incorporated into x-ray detector windows. The requirements for this application include high corrosion resistance, low intrinsic stress, high strains at failure, and high x-ray transmission. Low temperature sputtering was used because of its compatibility with the rest of the window fabrication process. Corrosion resistance was tested by exposure of carbon coated and uncoated Al thin films to humidity. Substrate curvature and bulge testing measurements were used to determine intrinsic stress and ultimate strain at failure. The composition and bonding of the carbon films were further characterized by electron energy loss spectroscopy, Raman spectroscopy, and carbon, hydrogen, and nitrogen elemental analyses. Samples had low compressive stress (down to.08 GPa), a high strain at failure (3%), and a low fraction of sp{sup 3} carbon–carbon bonds (less than 5%). The high breaking strain and excellent x-ray transmission of these sputtered carbon films indicate that they will work well as corrosion barriers in this application.

  12. X-ray radiometric analysis of lead and zinc concentrates using germanium radiation detector

    International Nuclear Information System (INIS)

    Vajgachev, A.A.; Mamysh, V.A.; Mil'chakov, V.I.; Shchekin, K.I.; Berezkin, V.V.

    1975-01-01

    The results of determination of lead, zinc and iron in lead and zinc concentrates by the X-ray-radiometric method with the use of germanium semiconductor detector are presented. In the experiments the 57 Co source and tritium-zirconium target were used. The activity of 57 Co was 2 mc. The area of the germanium detector employed was 5g mm 2 , its thickness - 2.3 mm. In lead concentrates zinc and iron were determined from the direct intensity of K-series radiation. In the analysis of zinc concentrates the same conditions of recording and excitation were used as in the case of lead concentrates, but the measurements were conducted in saturated layers. It is demonstrated that the use of germanium semiconductor detectors in combination with the suggested methods of measurements makes it possible to perform determination of iron, zinc and lead in zinc and lead concentrates with permissible error

  13. Analytical approximations to the Hotelling trace for digital x-ray detectors

    Science.gov (United States)

    Clarkson, Eric; Pineda, Angel R.; Barrett, Harrison H.

    2001-06-01

    The Hotelling trace is the signal-to-noise ratio for the ideal linear observer in a detection task. We provide an analytical approximation for this figure of merit when the signal is known exactly and the background is generated by a stationary random process, and the imaging system is an ideal digital x-ray detector. This approximation is based on assuming that the detector is infinite in extent. We test this approximation for finite-size detectors by comparing it to exact calculations using matrix inversion of the data covariance matrix. After verifying the validity of the approximation under a variety of circumstances, we use it to generate plots of the Hotelling trace as a function of pairs of parameters of the system, the signal and the background.

  14. Improvement of the instrumental line shape of X-ray spectrometers with Si(Li) - detectors

    International Nuclear Information System (INIS)

    Berdikov, V.V.; Zajtsev, E.A.; Iokhin, B.S.

    1983-01-01

    The possibility of decreasing the background of the X-ray spectrometer detector using the rise-time pulse selection method was investigated. Si(Li)-detectors of 10 and 25 mm 2 square were investigated. Spectrometer channel was composed of ORTEC-472 amplifier and ULTIMA/2 multichannel analyzer on the base of NOVA-3 minicomputer. The energy resolution was equal to 300 eV on 14 KeV line. The pulses of detection allowing were transmitted to analog-to-digital converter. The detection was allowed if front photopeak square) were measured at 17.4, 20.3 and 59.6 keV. 4-6-fold decrease of X-factor was obtained without any loss of detection efficiency. The combination of the method with collimation of radiation in the centre of the detector gives an extremely low value of X-factor which agress with theretical estimations

  15. A gamma- and X-ray detector for cryogenic, high magnetic field applications

    Energy Technology Data Exchange (ETDEWEB)

    Cooper, R.L., E-mail: roblcoop@indiana.edu [Indiana University, Bloomington, IN 47408 (United States); Alarcon, R. [Arizona State University, Tempe, AZ 85287 (United States); Bales, M.J. [University of Michigan, Ann Arbor, MI 48109 (United States); Bass, C.D. [National Institute of Standards and Technology, Stop 8461, NIST, Gaithersburg, MD 20899 (United States); Beise, E.J. [University of Maryland, College Park, MD 20742 (United States); Breuer, H., E-mail: breuer@enp.umd.edu [University of Maryland, College Park, MD 20742 (United States); Byrne, J. [University of Sussex, Brighton, BN1 9QH (United Kingdom); Chupp, T.E. [University of Michigan, Ann Arbor, MI 48109 (United States); Coakley, K.J. [National Institute of Standards and Technology, Boulder, CO 80305 (United States); Dewey, M.S.; Fu, C. [National Institute of Standards and Technology, Stop 8461, NIST, Gaithersburg, MD 20899 (United States); Gentile, T.R., E-mail: thomas.gentile@nist.gov [National Institute of Standards and Technology, Stop 8461, NIST, Gaithersburg, MD 20899 (United States); Mumm, H.P.; Nico, J.S. [National Institute of Standards and Technology, Stop 8461, NIST, Gaithersburg, MD 20899 (United States); O' Neill, B. [Arizona State University, Tempe, AZ 85287 (United States); Pulliam, K. [Tulane University, New Orleans, LA 70118 (United States); Thompson, A.K. [National Institute of Standards and Technology, Stop 8461, NIST, Gaithersburg, MD 20899 (United States); Wietfeldt, F.E. [Tulane University, New Orleans, LA 70118 (United States)

    2012-11-01

    As part of an experiment to measure the spectrum of photons emitted in beta-decay of the free neutron, we developed and operated a detector consisting of 12 bismuth germanate (BGO) crystals coupled to avalanche photodiodes (APDs). The detector was operated near liquid nitrogen temperature in the bore of a superconducting magnet and registered photons with energies from 5 keV to 1000 keV. To enlarge the detection range, we also directly detected soft X-rays with energies between 0.2 keV and 20 keV with three large area APDs. The construction and operation of the detector are presented, as well as information on operation of APDs at cryogenic temperatures.

  16. Characterization of spectrometric photon-counting X-ray detectors at different pitches

    International Nuclear Information System (INIS)

    Jurdit, M.; Moulin, V.; Ouvrier-Buffet, P.; Verger, L.; Brambilla, A.; Radisson, P.

    2017-01-01

    There is growing interest in energy-sensitive photon-counting detectors based on high flux X-ray imaging. Their potential applications include medical imaging, non-destructive testing and security. Innovative detectors of this type will need to count individual photons and sort them into selected energy bins, at several million counts per second and per mm 2 . Cd(Zn)Te detector grade materials with a thickness of 1.5 to 3 mm and pitches from 800 μm down to 200 μm were assembled onto interposer boards. These devices were tested using in-house-developed full-digital fast readout electronics. The 16-channel demonstrators, with 256 energy bins, were experimentally characterized by determining spectral resolution, count rate, and charge sharing, which becomes challenging at low pitch. Charge sharing correction was found to efficiently correct X-ray spectra up to 40 × 10 6 incident photons.s −1 .mm −2 .

  17. An X-ray imager based on silicon microstrip detector and coded mask

    International Nuclear Information System (INIS)

    Del Monte, E.; Costa, E.; Di Persio, G.; Donnarumma, I.; Evangelista, Y.; Feroci, M.; Frutti, M.; Lapshov, I.; Lazzarotto, F.; Mastropietro, M.; Morelli, E.; Pacciani, L.; Porrovecchio, G.; Rapisarda, M.; Rubini, A.; Soffitta, P.; Tavani, M.; Argan, A.

    2007-01-01

    SuperAGILE is the X-ray monitor of AGILE, a satellite mission for gamma-ray astronomy, and it is the first X-ray imaging instrument based on the technology of the silicon microstrip detectors combined with a coded aperture imaging technique. The SuperAGILE detection plane is composed of four 1-D silicon microstrip detector modules, mechanically coupled to tungsten coded mask units. The detector strips are separately and individually connected to the input analogue channels of the front-end electronics, composed of low-noise and low-power consumption VLSI ASIC chips. SuperAGILE can produce 1-D images with 6 arcmin angular resolution and ∼2-3 arcmin localisation capability, for intense sources, in a field of view composed of two orthogonal areas of 107 deg. x 68 deg. The time resolution is 2 μs, the overall dead time is ∼5 μs and the electronic noise is ∼7.5 keV full-width at half-maximum. The resulting instrument is very compact (40x40x14 cm 3 ), light (10 kg) and has low power consumption (12 W). AGILE is a mission of the Agenzia Spaziale Italiana and its launch is planned in 2007 in a low equatorial Earth orbit. In this contribution we present SuperAGILE and discuss its performance and scientific objectives

  18. Optimizing detector thickness in dual-shot dual-energy x-ray imaging

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Dong Woon; Kam, Soohwa; Youn, Hanbean; Kim, Ho Kyung [Pusan National University, Busan (Korea, Republic of)

    2015-05-15

    As a result, there exist apparent limitations in the conventional two-dimensional (2D) radiography: One is that the contrast between the structure of interest and the background in a radiograph is much less than the intrinsic subject contrast (i.e. the difference between their attenuation coefficients; Another is that the superimposed anatomical structures in the 2D radiograph results in an anatomical background clutter that may decrease the conspicuity of subtle underlying features. These limitations in spatial and material discrimination are important motivations for the recent development of 3D (e.g. tomosynthesis) and dual energy imaging (DEI) systems. DEI technique uses a combination of two images obtained at two different energies in successive x-ray exposures by rapidly switching the kilovolage (kV) applied to the x-ray tube. Commercial DEI systems usually employ a 'single' of flat-panel detector (FPD) to obtain two different kV images. However, we have a doubt in the use of the same detector for acquiring two different projections for the low- and high-kV setups because it is typically known that there exists an optimal detector thickness regarding specific imaging tasks or energies used.

  19. Monte-Carlo background simulations of present and future detectors in x-ray astronomy

    Science.gov (United States)

    Tenzer, C.; Kendziorra, E.; Santangelo, A.

    2008-07-01

    Reaching a low-level and well understood internal instrumental background is crucial for the scientific performance of an X-ray detector and, therefore, a main objective of the instrument designers. Monte-Carlo simulations of the physics processes and interactions taking place in a space-based X-ray detector as a result of its orbital environment can be applied to explain the measured background of existing missions. They are thus an excellent tool to predict and optimize the background of future observatories. Weak points of a design and the main sources of the background can be identified and methods to reduce them can be implemented and studied within the simulations. Using the Geant4 Monte-Carlo toolkit, we have created a simulation environment for space-based detectors and we present results of such background simulations for XMM-Newton's EPIC pn-CCD camera. The environment is also currently used to estimate and optimize the background of the future instruments Simbol-X and eRosita.

  20. Compton scattering artifacts in electron excited X-ray spectra measured with a silicon drift detector.

    Science.gov (United States)

    Ritchie, Nicholas W M; Newbury, Dale E; Lindstrom, Abigail P

    2011-12-01

    Artifacts are the nemesis of trace element analysis in electron-excited energy dispersive X-ray spectrometry. Peaks that result from nonideal behavior in the detector or sample can fool even an experienced microanalyst into believing that they have trace amounts of an element that is not present. Many artifacts, such as the Si escape peak, absorption edges, and coincidence peaks, can be traced to the detector. Others, such as secondary fluorescence peaks and scatter peaks, can be traced to the sample. We have identified a new sample-dependent artifact that we attribute to Compton scattering of energetic X-rays generated in a small feature and subsequently scattered from a low atomic number matrix. It seems likely that this artifact has not previously been reported because it only occurs under specific conditions and represents a relatively small signal. However, with the advent of silicon drift detectors and their utility for trace element analysis, we anticipate that more people will observe it and possibly misidentify it. Though small, the artifact is not inconsequential. Under some conditions, it is possible to mistakenly identify the Compton scatter artifact as approximately 1% of an element that is not present.

  1. Low-energy X-ray detection in cryogenic detectors with tungsten thermometers

    International Nuclear Information System (INIS)

    Colling, P.; Nucciotti, A.; Bucci, C.; Cooper, S.; Ferger, P.; Frank, M.; Nagel, U.; Proebst, F.; Seidel, W.

    1994-08-01

    In the course of our development of calorimetric particle detectors with superconducting phase transition thermometers, we have succeeded in depositing epitaxial α-tungsten films on sapphire which have critical temperatures T c near 15 mK. To our knowledge this is the first time that the T c of bulk tungsten has been observed in thin films. Such films used as thermometers are very sensitive and provide good energy resolution: with 4 g and 32 g sapphire crystals energy resolutions of better than 100eV (FWHM) for 1.5 KeV X-rays have been achieved. (orig.)

  2. Characterization of an x-ray hybrid CMOS detector with low interpixel capacitive crosstalk

    OpenAIRE

    Griffith, Christopher V.; Bongiorno, Stephen D.; Burrows, David N.; Falcone, Abraham D.; Prieskorn, Zachary R.

    2012-01-01

    We present the results of x-ray measurements on a hybrid CMOS detector that uses a H2RG ROIC and a unique bonding structure. The silicon absorber array has a 36{\\mu}m pixel size, and the readout array has a pitch of 18{\\mu}m; but only one readout circuit line is bonded to each 36x36{\\mu}m absorber pixel. This unique bonding structure gives the readout an effective pitch of 36{\\mu}m. We find the increased pitch between readout bonds significantly reduces the interpixel capacitance of the CMOS ...

  3. Mobility-lifetime product in epitaxial GaAs X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Sun, G.C. [GESEC R and D, Universite Pierre et Marie Curie, Bat.11, 140 rue de Lourmel, 75015 Paris (France)]. E-mail: guocsun@ccr.jussieu.fr; Zazoui, M. [LPMC, Faculte des Sciences et Techniques-Mohammedia, B.P. 146 Bd Hassan II, Mohammedia, Maroc (Morocco); Talbi, N. [Faculte des Sciences, Universite de Gabes, Route de Medenine, 6029 Gabes (Tunisia); Khirouni, K. [Faculte des Sciences, Universite de Gabes, Route de Medenine, 6029 Gabes (Tunisia); Bourgoin, J.C. [GESEC R and D, Universite Pierre et Marie Curie, Bat.11, 140 rue de Lourmel, 75015 Paris (France)

    2007-04-01

    Self-supported thick (200-500 {mu}m), non-intentionally doped, epitaxial GaAs layers are good candidates for X-ray imaging for the following reasons. Their electronic properties are homogeneous over large areas, they can be grown at low cost, the technology to realize pixel detectors of various size is standard, the defect concentration is low and the fluorescence yield is small. Here, we characterize the defects present in the material and evaluate the mobility-lifetime product, using Deep Level Transient Spectroscopy combined with current-voltage and charge collection measurements.

  4. The data acquisition system for the Leeds Infirmary MWPC X-ray imaging detector

    International Nuclear Information System (INIS)

    Quinton, S.; Gibbings, D.; Jones, D.; Norton, H.

    1979-10-01

    An electronic system is described which is designed to acquire and process data from a MWPC X-ray imaging detector. Two dimensional information from the chamber is obtained by using cathode plane delay-line readout. A single crate CAMAC assembly is used as the chamber-computer interface. The use of control source units for the delay line scalers and TV display driver functions together with an intermediate memory in the crate allows input data rates up to 1MHz and TV display facilities without constant computer refreshing. (author)

  5. X-ray measurement with Pin type semiconductor detectors; Medicion de rayos X con detectores de semiconductor tipo PIN

    Energy Technology Data Exchange (ETDEWEB)

    Ramirez J, F.J. [Instituto Nacional de Investigaciones Nucleares, Departamento de Electronica, C.P. 52045 Salazar, Estado de Mexico (Mexico)

    2000-07-01

    Here are presented the experimental results of the applications of Pin type radiation detectors developed in a National Institute of Nuclear Research (ININ) project, in the measurement of low energy gamma and X-rays. The applications were oriented mainly toward the Medical Physics area. It is planned other applications which are in process of implementation inside the National Institute of Nuclear Research in Mexico. (Author)

  6. Improvement of the detector resolution in X-ray spectrometry by using the maximum entropy method

    International Nuclear Information System (INIS)

    Fernández, Jorge E.; Scot, Viviana; Giulio, Eugenio Di; Sabbatucci, Lorenzo

    2015-01-01

    In every X-ray spectroscopy measurement the influence of the detection system causes loss of information. Different mechanisms contribute to form the so-called detector response function (DRF): the detector efficiency, the escape of photons as a consequence of photoelectric or scattering interactions, the spectrum smearing due to the energy resolution, and, in solid states detectors (SSD), the charge collection artifacts. To recover the original spectrum, it is necessary to remove the detector influence by solving the so-called inverse problem. The maximum entropy unfolding technique solves this problem by imposing a set of constraints, taking advantage of the known a priori information and preserving the positive-defined character of the X-ray spectrum. This method has been included in the tool UMESTRAT (Unfolding Maximum Entropy STRATegy), which adopts a semi-automatic strategy to solve the unfolding problem based on a suitable combination of the codes MAXED and GRAVEL, developed at PTB. In the past UMESTRAT proved the capability to resolve characteristic peaks which were revealed as overlapped by a Si SSD, giving good qualitative results. In order to obtain quantitative results, UMESTRAT has been modified to include the additional constraint of the total number of photons of the spectrum, which can be easily determined by inverting the diagonal efficiency matrix. The features of the improved code are illustrated with some examples of unfolding from three commonly used SSD like Si, Ge, and CdTe. The quantitative unfolding can be considered as a software improvement of the detector resolution. - Highlights: • Radiation detection introduces distortions in X- and Gamma-ray spectrum measurements. • UMESTRAT is a graphical tool to unfold X- and Gamma-ray spectra. • UMESTRAT uses the maximum entropy method. • UMESTRAT’s new version produces unfolded spectra with quantitative meaning. • UMESTRAT is a software tool to improve the detector resolution.

  7. Multi-energy x-ray detectors to improve air-cargo security

    Science.gov (United States)

    Paulus, Caroline; Moulin, Vincent; Perion, Didier; Radisson, Patrick; Verger, Loïck

    2017-05-01

    X-ray based systems have been used for decades to screen luggage or cargo to detect illicit material. The advent of energy-sensitive photon-counting x-ray detectors mainly based on Cd(Zn)Te semi-conductor technology enables to improve discrimination between materials compared to single or dual energy technology. The presented work is part of the EUROSKY European project to develop a Single European Secure Air-Cargo Space. "Cargo" context implies the presence of relatively heavy objects and with potentially high atomic number. All the study is conducted on simulations with three different detectors: a typical dual energy sandwich detector, a realistic model of the commercial ME100 multi-energy detector marketed by MULTIX, and a ME100 "Cargo": a not yet existing modified multi-energy version of the ME100 more suited to air freight cargo inspection. Firstly, a comparison on simulated measurements shows the performances improvement of the new multi-energy detectors compared to the current dual-energy one. The relative performances are evaluated according to different criteria of separability or contrast-to-noise ratio and the impact of different parameters is studied (influence of channel number, type of materials and tube voltage). Secondly, performances of multi-energy detectors for overlaps processing in a dual-view system is accessed: the case of orthogonal projections has been studied, one giving dimensional values, the other one providing spectral data to assess effective atomic number. A method of overlap correction has been proposed and extended to multi-layer objects case. Therefore, Calibration and processing based on bi-material decomposition have been adapted for this purpose.

  8. Investigation of chemical vapour deposition diamond detectors by X-ray micro-beam induced current and X-ray micro-beam induced luminescence techniques

    International Nuclear Information System (INIS)

    Olivero, P.; Manfredotti, C.; Vittone, E.; Fizzotti, F.; Paolini, C.; Lo Giudice, A.; Barrett, R.; Tucoulou, R.

    2004-01-01

    Tracking detectors have become an important ingredient in high-energy physics experiments. In order to survive the harsh detection environment of the large hadron collider (LHC), trackers need to have special properties. They must be radiation hard, provide fast collection of charge, be as thin as possible and remove heat from readout electronics. The unique properties of diamond allow it to fulfill these requirements. In this work we present an investigation of the charge transport and luminescence properties of 'detector grade' artificial chemical vapour deposition (CVD) diamond devices developed within the CERN RD42 collaboration, performed by means of X-ray micro-beam induced current collection (XBICC) and X-ray micro-beam induced luminescence (XBIL) techniques. XBICC technique allows quantitative estimates of the transport parameters of the material to be evaluated and mapped with micrometric spatial resolution. In particular, the high resolution and sensitivity of the technique has allowed a quantitative study of the inhomogeneity of the charge transport parameter defined as the product of mobility and lifetime for both electron and holes. XBIL represents a technique complementary to ion beam induced luminescence (IBIL), which has already been used by our group, since X-ray energy loss profile in the material is different from that of MeV ions. X-ray induced luminescence maps have been performed simultaneously with induced photocurrent maps, to correlate charge transport and induced luminescence properties of diamond. Simultaneous XBICC and XBIL maps exhibit features of partial complementarity that have been interpreted on the basis of considerations on radiative and non-radiative recombination processes which compete with charge transport efficiency

  9. One dimensional detector for X-ray diffraction with superior energy resolution based on silicon strip detector technology

    International Nuclear Information System (INIS)

    Dąbrowski, W; Fiutowski, T; Wiącek, P; Fink, J; Krane, H-G

    2012-01-01

    1-D position sensitive X-ray detectors based on silicon strip detector technology have become standard instruments in X-ray diffraction and are available from several vendors. As these devices have been proven to be very useful and efficient further improvement of their performance is investigated. The silicon strip detectors in X-ray diffraction are primarily used as counting devices and the requirements concerning the spatial resolution, dynamic range and count rate capability are of primary importance. However, there are several experimental issues in which a good energy resolution is important. The energy resolution of silicon strip detectors is limited by the charge sharing effects in the sensor as well as by noise of the front-end electronics. The charge sharing effects in the sensor and various aspects of the electronics, including the baseline fluctuations, which affect the energy resolution, have been analyzed in detail and a new readout concept has been developed. A front-end ASIC with a novel scheme of baseline restoration and novel interstrip logic circuitry has been designed. The interstrip logic is used to reject the events resulting in significant charge sharing between neighboring strips. At the expense of rejecting small fraction of photons entering the detector one can obtain single strip energy spectra almost free of charge sharing effects. In the paper we present the design considerations and measured performance of the detector being developed. The electronic noise of the system at room temperature is typically of the order of 70 el rms for 17 mm long silicon strips and a peaking time of about 1 μs. The energy resolution of 600 eV FWHM has been achieved including the non-reducible charge sharing effects and the electronic noise. This energy resolution is sufficient to address a common problem in X-ray diffraction, i.e. electronic suppression of the fluorescence radiation from samples containing iron or cobalt while irradiated with 8.04 ke

  10. Heel effect adaptive flat field correction of digital x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Yu, Yongjian [X-ray Products, Varian Medical Systems Inc., Liverpool, New York 13088 (United States); Wang, Jue [Department of Mathematics, Union College, Schenectady, New York 12308 (United States)

    2013-08-15

    Purpose: Anode heel effect renders large-scale background nonuniformities in digital radiographs. Conventional offset/gain calibration is performed at mono source-to-image distance (SID), and disregards the SID-dependent characteristic of heel effect. It results in a residual nonuniform background in the corrected radiographs when the SID settings for calibration and correction differ. In this work, the authors develop a robust and efficient computational method for digital x-ray detector gain correction adapted to SID-variant heel effect, without resorting to physical filters, phantoms, complicated heel effect models, or multiple-SID calibration and interpolation.Methods: The authors present the Duo-SID projection correction method. In our approach, conventional offset/gain calibrations are performed only twice, at the minimum and maximum SIDs of the system in typical clinical use. A fast iterative separation algorithm is devised to extract the detector gain and basis heel patterns from the min/max SID calibrations. The resultant detector gain is independent of SID, while the basis heel patterns are parameterized by the min- and max-SID. The heel pattern at any SID is obtained from the min-SID basis heel pattern via projection imaging principles. The system gain desired at a specific acquisition SID is then constructed using the projected heel pattern and detector gain map.Results: The method was evaluated for flat field and anatomical phantom image corrections. It demonstrated promising improvements over interpolation and conventional gain calibration/correction methods, lowering their correction errors by approximately 70% and 80%, respectively. The separation algorithm was able to extract the detector gain and heel patterns with less than 2% error, and the Duo-SID corrected images showed perceptually appealing uniform background across the detector.Conclusions: The Duo-SID correction method has substantially improved on conventional offset/gain corrections for

  11. Heel effect adaptive flat field correction of digital x-ray detectors

    International Nuclear Information System (INIS)

    Yu, Yongjian; Wang, Jue

    2013-01-01

    Purpose: Anode heel effect renders large-scale background nonuniformities in digital radiographs. Conventional offset/gain calibration is performed at mono source-to-image distance (SID), and disregards the SID-dependent characteristic of heel effect. It results in a residual nonuniform background in the corrected radiographs when the SID settings for calibration and correction differ. In this work, the authors develop a robust and efficient computational method for digital x-ray detector gain correction adapted to SID-variant heel effect, without resorting to physical filters, phantoms, complicated heel effect models, or multiple-SID calibration and interpolation.Methods: The authors present the Duo-SID projection correction method. In our approach, conventional offset/gain calibrations are performed only twice, at the minimum and maximum SIDs of the system in typical clinical use. A fast iterative separation algorithm is devised to extract the detector gain and basis heel patterns from the min/max SID calibrations. The resultant detector gain is independent of SID, while the basis heel patterns are parameterized by the min- and max-SID. The heel pattern at any SID is obtained from the min-SID basis heel pattern via projection imaging principles. The system gain desired at a specific acquisition SID is then constructed using the projected heel pattern and detector gain map.Results: The method was evaluated for flat field and anatomical phantom image corrections. It demonstrated promising improvements over interpolation and conventional gain calibration/correction methods, lowering their correction errors by approximately 70% and 80%, respectively. The separation algorithm was able to extract the detector gain and heel patterns with less than 2% error, and the Duo-SID corrected images showed perceptually appealing uniform background across the detector.Conclusions: The Duo-SID correction method has substantially improved on conventional offset/gain corrections for

  12. X-ray fluorescence analysis in environmental radiological surveillance using HPGe detectors

    International Nuclear Information System (INIS)

    Herrera Peraza, E.; Renteria Villalobos, M.; Montero Cabrera, M.E.; Munoz Romero, A.

    2004-01-01

    X-ray fluorescence (XRF) has been proven to be a valuable tool for determining trace quantities of heavy metals, such as uranium and lead, in different types of samples. The present paper demonstrates the applicability of XRF spectrometry to measure the concentrations of these heavy metals in samples from natural ore and soil. The values of uranium concentrations in rock from the Pena Blanca uranium ore, in Chihuahua, Mexico, were calculated for the purpose of precertifying the rock powders samples. The comparison with other techniques, such as inductively coupled plasma atomic emission spectrometry, atomic absorption spectrometry, alpha spectrometry and electron microscopy, was used to complete the precertification process, so that the sample powders may be used as secondary standards. The source-sample-detector geometry and the incident angle are the most important factors for obtaining low detection limits. The selected system uses a 57 Co source of about 0.1 mCi to excite the K X-rays from uranium and lead. X-rays were recorded on a CANBERRA HPGe coaxial detector. The comparative results for two incident angles (90 deg and 180 deg ) performed previously by other authors show that the best geometry is the backscattering geometry. In the present paper, using EGS4 code system with Monte Carlo simulation, it was possible to determine the location and distribution of background produced by the Compton edge in the optimized geometry. This procedure allowed to find the minimum detectable concentration of uranium and lead, which was experimentally calculated using standards. The possibility of performing in vivo measurements rapidly and easily, as well as the factors affecting accuracy and the minimum detectable concentration in several samples are also discussed

  13. X-ray fluorescence analysis in environmental radiological surveillance using HPGe detectors

    Energy Technology Data Exchange (ETDEWEB)

    Herrera Peraza, E. [Department of Environmental Radiological Surveillance, Centro de Investigacion en Materiales Avanzados (CIMAV), P.O. Box 31109, Miguel de Cervantes no. 120, Complejo Industrial Chihuahua, Chihuahua (Mexico)]. E-mail: eduardo.herrera@cimav.edu.mx; Renteria Villalobos, M. [Department of Environmental Radiological Surveillance, Centro de Investigacion en Materiales Avanzados (CIMAV), P.O. Box 31109, Miguel de Cervantes no. 120, Complejo Industrial Chihuahua, Chihuahua (Mexico); Montero Cabrera, M.E. [Department of Environmental Radiological Surveillance, Centro de Investigacion en Materiales Avanzados (CIMAV), P.O. Box 31109, Miguel de Cervantes no. 120, Complejo Industrial Chihuahua, Chihuahua (Mexico); Munoz Romero, A. [Department of Environmental Radiological Surveillance, Centro de Investigacion en Materiales Avanzados (CIMAV), P.O. Box 31109, Miguel de Cervantes no. 120, Complejo Industrial Chihuahua, Chihuahua (Mexico)

    2004-10-08

    X-ray fluorescence (XRF) has been proven to be a valuable tool for determining trace quantities of heavy metals, such as uranium and lead, in different types of samples. The present paper demonstrates the applicability of XRF spectrometry to measure the concentrations of these heavy metals in samples from natural ore and soil. The values of uranium concentrations in rock from the Pena Blanca uranium ore, in Chihuahua, Mexico, were calculated for the purpose of precertifying the rock powders samples. The comparison with other techniques, such as inductively coupled plasma atomic emission spectrometry, atomic absorption spectrometry, alpha spectrometry and electron microscopy, was used to complete the precertification process, so that the sample powders may be used as secondary standards. The source-sample-detector geometry and the incident angle are the most important factors for obtaining low detection limits. The selected system uses a {sup 57}Co source of about 0.1 mCi to excite the K X-rays from uranium and lead. X-rays were recorded on a CANBERRA HPGe coaxial detector. The comparative results for two incident angles (90 deg and 180 deg ) performed previously by other authors show that the best geometry is the backscattering geometry. In the present paper, using EGS4 code system with Monte Carlo simulation, it was possible to determine the location and distribution of background produced by the Compton edge in the optimized geometry. This procedure allowed to find the minimum detectable concentration of uranium and lead, which was experimentally calculated using standards. The possibility of performing in vivo measurements rapidly and easily, as well as the factors affecting accuracy and the minimum detectable concentration in several samples are also discussed.

  14. A VXI-based high speed x-ray CCD detector

    International Nuclear Information System (INIS)

    Huang, Qiang; Hopf, R.; Rodricks, B.

    1993-01-01

    For time-resolved x-ray scattering, one ideally wants a high speed detector that also is capable of giving position sensitive information. Charge Coupled Devices (CCDS) have been used successfully as x-ray detectors. Unfortunately, they are inherently slow because of the serial readout EEV has developed a CCD that has eight channels of parallel readout, thus increasing the speed eight fold. Using state-of-the-art VXI electronics, we have developed a readout system that could read the entire array in 2.5 ms using a 20-MHz readout clock. For testing and characterization the device was clocked at a significantly slower speed of 30 kHz. The data is preamplified and all eight channels of output are simultaneously digitized to 12 bits and stored in buffer memory. The system is controlled by a 486-based PC through an MXI bus and VXI controller using commercially available software. The system is also capable of real-time image display and manipulation

  15. P-type silicon surface barrier detector used for x-ray dosimetry

    International Nuclear Information System (INIS)

    Yamamoto, Hisao; Hatakeyama, Satoru; Norimura, Toshiyuki; Tsuchiya, Takehiko

    1983-01-01

    Responses to X-rays of a P-type surface barrier detector fabricated in our laboratory were studied, taking into consideration the dependence on the temperature in order to examine its applicability to dosimetry of short-range radiation. The study was also made in the case of N-type surface barrier detector. At room temperature, the short-circuit current increased linearly with exposure dose rate (15 - 50 R/min) for N- and P-type detectors. The open-circuit voltage showed a nonlinear dependence. With increasing temperature, the short-circuit current for the N-type detector was approximately constant up to 30 0 C and then decreased, though the open-circuit voltage decreased linearly. For the P- type detector, both open-circuit voltage and short-circuit current decreased almost linearly with increasing temperature. While a P-type detector is still open to some improvements, these results indicate that it can be used as a dosimeter. (author)

  16. Energy Reconstruction for Events Detected in TES X-ray Detectors

    Science.gov (United States)

    Ceballos, M. T.; Cardiel, N.; Cobo, B.

    2015-09-01

    The processing of the X-ray events detected by a TES (Transition Edge Sensor) device (such as the one that will be proposed in the ESA AO call for instruments for the Athena mission (Nandra et al. 2013) as a high spectral resolution instrument, X-IFU (Barret et al. 2013)), is a several step procedure that starts with the detection of the current pulses in a noisy signal and ends up with their energy reconstruction. For this last stage, an energy calibration process is required to convert the pseudo energies measured in the detector to the real energies of the incoming photons, accounting for possible nonlinearity effects in the detector. We present the details of the energy calibration algorithm we implemented as the last part of the Event Processing software that we are developing for the X-IFU instrument, that permits the calculation of the calibration constants in an analytical way.

  17. Design and image-quality performance of high resolution CMOS-based X-ray imaging detectors for digital mammography

    Science.gov (United States)

    Cha, B. K.; Kim, J. Y.; Kim, Y. J.; Yun, S.; Cho, G.; Kim, H. K.; Seo, C.-W.; Jeon, S.; Huh, Y.

    2012-04-01

    In digital X-ray imaging systems, X-ray imaging detectors based on scintillating screens with electronic devices such as charge-coupled devices (CCDs), thin-film transistors (TFT), complementary metal oxide semiconductor (CMOS) flat panel imagers have been introduced for general radiography, dental, mammography and non-destructive testing (NDT) applications. Recently, a large-area CMOS active-pixel sensor (APS) in combination with scintillation films has been widely used in a variety of digital X-ray imaging applications. We employed a scintillator-based CMOS APS image sensor for high-resolution mammography. In this work, both powder-type Gd2O2S:Tb and a columnar structured CsI:Tl scintillation screens with various thicknesses were fabricated and used as materials to convert X-ray into visible light. These scintillating screens were directly coupled to a CMOS flat panel imager with a 25 × 50 mm2 active area and a 48 μm pixel pitch for high spatial resolution acquisition. We used a W/Al mammographic X-ray source with a 30 kVp energy condition. The imaging characterization of the X-ray detector was measured and analyzed in terms of linearity in incident X-ray dose, modulation transfer function (MTF), noise-power spectrum (NPS) and detective quantum efficiency (DQE).

  18. Improvement of density resolution in short-pulse hard x-ray radiographic imaging using detector stacks

    Energy Technology Data Exchange (ETDEWEB)

    Borm, B.; Gärtner, F.; Khaghani, D. [GSI Helmholtzzentrum für Schwerionenforschung, Darmstadt (Germany); Johann Wolfgang Goethe-Universität, Frankfurt am Main (Germany); Neumayer, P. [GSI Helmholtzzentrum für Schwerionenforschung, Darmstadt (Germany)

    2016-09-15

    We demonstrate that stacking several imaging plates (IPs) constitutes an easy method to increase hard x-ray detection efficiency. Used to record x-ray radiographic images produced by an intense-laser driven hard x-ray backlighter source, the IP stacks resulted in a significant improvement of the radiograph density resolution. We attribute this to the higher quantum efficiency of the combined detectors, leading to a reduced photon noise. Electron-photon transport simulations of the interaction processes in the detector reproduce the observed contrast improvement. Increasing the detection efficiency to enhance radiographic imaging capabilities is equally effective as increasing the x-ray source yield, e.g., by a larger drive laser energy.

  19. Taheri-Saramad x-ray detector (TSXD): a novel high spatial resolution x-ray imager based on ZnO nano scintillator wires in polycarbonate membrane.

    Science.gov (United States)

    Taheri, A; Saramad, S; Ghalenoei, S; Setayeshi, S

    2014-01-01

    A novel x-ray imager based on ZnO nanowires is designed and fabricated. The proposed architecture is based on scintillation properties of ZnO nanostructures in a polycarbonate track-etched membrane. Because of higher refractive index of ZnO nanowire compared to the membrane, the nanowire acts as an optical fiber that prevents the generated optical photons to spread inside the detector. This effect improves the spatial resolution of the imager. The detection quantum efficiency and spatial resolution of the fabricated imager are 11% and <6.8 μm, respectively.

  20. Taheri-Saramad x-ray detector (TSXD): A novel high spatial resolution x-ray imager based on ZnO nano scintillator wires in polycarbonate membrane

    Energy Technology Data Exchange (ETDEWEB)

    Taheri, A., E-mail: at1361@aut.ac.ir; Saramad, S.; Ghalenoei, S.; Setayeshi, S. [Department of Energy Engineering and Physics, Amirkabir University of Technology, Tehran 15875-4413 (Iran, Islamic Republic of)

    2014-01-15

    A novel x-ray imager based on ZnO nanowires is designed and fabricated. The proposed architecture is based on scintillation properties of ZnO nanostructures in a polycarbonate track-etched membrane. Because of higher refractive index of ZnO nanowire compared to the membrane, the nanowire acts as an optical fiber that prevents the generated optical photons to spread inside the detector. This effect improves the spatial resolution of the imager. The detection quantum efficiency and spatial resolution of the fabricated imager are 11% and <6.8 μm, respectively.

  1. Static and time-resolved 10-1000 keV x-ray imaging detector options for NIF

    International Nuclear Information System (INIS)

    Landen, O.L.; Bell, P.M.; McDonald, J.W.; Park, H.-S.; Weber, F.; Moody, J.D.; Lowry, M.E.; Stewart, R.E.

    2004-01-01

    High energy (>10 keV) x-ray self-emission imaging and radiography will be essential components of many NIF high energy density physics experiments. In preparation for such experiments, we have evaluated the pros and cons of various static [x-ray film, bare charge-coupled device (CCD), and scintillator + CCD] and time-resolved (streaked and gated) 10-1000 keV detectors

  2. A high resolution, high counting rate bidimensional, MWPC imaging detector for small angle X-ray diffraction studies

    International Nuclear Information System (INIS)

    Bateman, J.E.; Connolly, J.F.; Sawyer, E.C.; Stephenson, R.

    1981-07-01

    The performance is reported of a 200 mm x 200 mm X-ray imaging MWPC aimed at applications in small angle X-ray diffraction and scattering. With quantum energies of approximately 8 keV high spatial resolution (+- 0.5 mm x +- 0.14 mm) with a capability for data taking at >approximately 350 kHz is reported. The detection efficiency is approximately 75% and the detector operates as a sealed unit with a long lifetime. (author)

  3. Spectral and spatial resolution properties of photon counting X-ray detectors like the Medipix-Detector

    International Nuclear Information System (INIS)

    Korn, A.

    2007-01-01

    The Medipix detector is a hybrid photon counting X-ray detector, consisting of an ASIC and a semiconducting layer as the sensor. This makes the Medipix a direct converting detector. A special feature of the Medipix is a signal processing circuit in every single pixel. This circuit amplifies the input signal triggered by a photon and then transforms the pulse into a digital signal. This early stage digitalisation is one of the main advantages of the detector, since no dark currents are integrated into the signal. Furthermore, the energy information of each single photon is partly preserved. The high number of pixels lends the detector a wide dynamic range, starting from single counts up to a rate of 1010 photons per cm2 and second. Apart from the many advantages, there are still some problems with the detector. Some effects lead to a deterioration of the energy resolution as well as the spatial resolution. The main reasons for this are two effects occuring in the detector, charge sharing and backscattering inside the detector. This study investigates the influence of those two effects on both the energy and spatial resolution. The physical causes of these effects are delineated and their impact on the detector output is examined. In contrast to high energy photon detectors, the repulsion of the charge carriers drifting inside the sensor must not be neglected in a detailed model of X-ray detectors with an energy range of 5 keV-200 keV. For the simulation of the Medipix using Monte Carlo simulations, the software ROSI was augmented. The added features allow a detailed simulation of the charge distribution, using the relevant physical effects that alter the distribution width during the drift towards the sensor electrodes as well further influences on the detector output, including electronical noise, threshold noise or the geometry of the detector. The measured energy and spatial resolution of several different models of Medipix is compared to the simulated

  4. X-ray Hybrid CMOS Detectors : Recent progress in development and characterization

    Science.gov (United States)

    Chattopadhyay, Tanmoy; Falcone, Abraham; Burrows, David N.

    2017-08-01

    PennState high energy astronomy laboratory has been working on the development and characterization of Hybrid CMOS Detectors (HCDs) for last few years in collaboration with Teledyne Imaging Sensors (TIS). HCDs are preferred over X-ray CCDs due to their higher and flexible read out rate, radiation hardness and low power which make them more suitable for next generation large area X-ray telescopic missions. An H2RG detector with 36 micron pixel pitch and 18 micron ROIC, has been selected for a sounding rocket flight in 2018. The H2RG detector provides ~2.5 % energy resolution at 5.9 keV and ~7 e- read noise when coupled to a cryo-SIDECAR. We could also detect a clear Oxygen line (~0.5 keV) from the detector implying a lower energy threshold of ~0.3 keV. Further improvement in the energy resolution and read noise is currently under progress. We have been working on the characterization of small pixel HCDs (12.5 micron pixel; smallest pixel HCDs developed so far) which is important for the development of next generation high resolution X-ray spectroscopic instrument based on HCDs. Event recognition in HCDs is another exciting prospect which have been successfully shown to work with a 64 X 64 pixel prototype SPEEDSTAR-EXD which use comparators at each pixel to read out only those pixels having detectable signal, thereby providing an order of magnitude improvement in the read out rate. Currently, we are working on the development of a large area SPEEDSTAR-EXD array for the development of a full fledged instrument. HCDs due to their fast read out, can also be explored as a large FOV instrument to study GRB afterglows and variability and spectroscopic study of other astrophysical transients. In this context, we are characterizing a Lobster-HCD system at multiple energies and multiple off-axis angles for future rocket or CubeSate experiments. In this presentation, I will briefly present these new developments and experiments with HCDs and the analysis techniques.

  5. Basis material decomposition method for material discrimination with a new spectrometric X-ray imaging detector

    Science.gov (United States)

    Brambilla, A.; Gorecki, A.; Potop, A.; Paulus, C.; Verger, L.

    2017-08-01

    Energy sensitive photon counting X-ray detectors provide energy dependent information which can be exploited for material identification. The attenuation of an X-ray beam as a function of energy depends on the effective atomic number Zeff and the density. However, the measured attenuation is degraded by the imperfections of the detector response such as charge sharing or pile-up. These imperfections lead to non-linearities that limit the benefits of energy resolved imaging. This work aims to implement a basis material decomposition method which overcomes these problems. Basis material decomposition is based on the fact that the attenuation of any material or complex object can be accurately reproduced by a combination of equivalent thicknesses of basis materials. Our method is based on a calibration phase to learn the response of the detector for different combinations of thicknesses of the basis materials. The decomposition algorithm finds the thicknesses of basis material whose spectrum is closest to the measurement, using a maximum likelihood criterion assuming a Poisson law distribution of photon counts for each energy bin. The method was used with a ME100 linear array spectrometric X-ray imager to decompose different plastic materials on a Polyethylene and Polyvinyl Chloride base. The resulting equivalent thicknesses were used to estimate the effective atomic number Zeff. The results are in good agreement with the theoretical Zeff, regardless of the plastic sample thickness. The linear behaviour of the equivalent lengths makes it possible to process overlapped materials. Moreover, the method was tested with a 3 materials base by adding gadolinium, whose K-edge is not taken into account by the other two materials. The proposed method has the advantage that it can be used with any number of energy channels, taking full advantage of the high energy resolution of the ME100 detector. Although in principle two channels are sufficient, experimental measurements show

  6. Polarization phenomena in Al/p-CdTe/Pt X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Principato, F., E-mail: fabio.principato@unipa.it; Turturici, A.A.; Gallo, M.; Abbene, L.

    2013-12-01

    Over the last decades, CdTe detectors are widely used for the development of room temperature X-ray and gamma ray spectrometers. Typically, high resolution CdTe detectors are fabricated with blocking contacts (indium and aluminum) ensuring low leakage currents and high electric field for optimum charge collection. As well known, time instability under bias voltage (termed as polarization) is the major drawback of CdTe diode detectors. Polarization phenomena cause a progressive degradation of the spectroscopic performance with time, due to hole trapping and detrapping from deep acceptors levels. In this work, we studied the polarization phenomenon on new Al/p-CdTe/Pt detectors, manufactured by Acrorad (Japan), through electrical and spectroscopic approaches. In particular, we investigated on the time degradation of the spectroscopic response of the detectors at different temperatures, voltages and energies. Current transient measurements were also performed to better understand the properties of the deep acceptor levels and their correlation with the polarization effect.

  7. Graphical user interface for a dual-module EMCCD x-ray detector array

    Science.gov (United States)

    Wang, Weiyuan; Ionita, Ciprian; Kuhls-Gilcrist, Andrew; Huang, Ying; Qu, Bin; Gupta, Sandesh K.; Bednarek, Daniel R.; Rudin, Stephen

    2011-03-01

    A new Graphical User Interface (GUI) was developed using Laboratory Virtual Instrumentation Engineering Workbench (LabVIEW) for a high-resolution, high-sensitivity Solid State X-ray Image Intensifier (SSXII), which is a new x-ray detector for radiographic and fluoroscopic imaging, consisting of an array of Electron-Multiplying CCDs (EMCCDs) each having a variable on-chip electron-multiplication gain of up to 2000x to reduce the effect of readout noise. To enlarge the field-of-view (FOV), each EMCCD sensor is coupled to an x-ray phosphor through a fiberoptic taper. Two EMCCD camera modules are used in our prototype to form a computer-controlled array; however, larger arrays are under development. The new GUI provides patient registration, EMCCD module control, image acquisition, and patient image review. Images from the array are stitched into a 2kx1k pixel image that can be acquired and saved at a rate of 17 Hz (faster with pixel binning). When reviewing the patient's data, the operator can select images from the patient's directory tree listed by the GUI and cycle through the images using a slider bar. Commonly used camera parameters including exposure time, trigger mode, and individual EMCCD gain can be easily adjusted using the GUI. The GUI is designed to accommodate expansion of the EMCCD array to even larger FOVs with more modules. The high-resolution, high-sensitivity EMCCD modular-array SSXII imager with the new user-friendly GUI should enable angiographers and interventionalists to visualize smaller vessels and endovascular devices, helping them to make more accurate diagnoses and to perform more precise image-guided interventions.

  8. Graphical User Interface for a Dual-Module EMCCD X-ray Detector Array.

    Science.gov (United States)

    Wang, Weiyuan; Ionita, Ciprian; Kuhls-Gilcrist, Andrew; Huang, Ying; Qu, Bin; Gupta, Sandesh K; Bednarek, Daniel R; Rudin, Stephen

    2011-03-16

    A new Graphical User Interface (GUI) was developed using Laboratory Virtual Instrumentation Engineering Workbench (LabVIEW) for a high-resolution, high-sensitivity Solid State X-ray Image Intensifier (SSXII), which is a new x-ray detector for radiographic and fluoroscopic imaging, consisting of an array of Electron-Multiplying CCDs (EMCCDs) each having a variable on-chip electron-multiplication gain of up to 2000× to reduce the effect of readout noise. To enlarge the field-of-view (FOV), each EMCCD sensor is coupled to an x-ray phosphor through a fiberoptic taper. Two EMCCD camera modules are used in our prototype to form a computer-controlled array; however, larger arrays are under development. The new GUI provides patient registration, EMCCD module control, image acquisition, and patient image review. Images from the array are stitched into a 2k×1k pixel image that can be acquired and saved at a rate of 17 Hz (faster with pixel binning). When reviewing the patient's data, the operator can select images from the patient's directory tree listed by the GUI and cycle through the images using a slider bar. Commonly used camera parameters including exposure time, trigger mode, and individual EMCCD gain can be easily adjusted using the GUI. The GUI is designed to accommodate expansion of the EMCCD array to even larger FOVs with more modules. The high-resolution, high-sensitivity EMCCD modular-array SSXII imager with the new user-friendly GUI should enable angiographers and interventionalists to visualize smaller vessels and endovascular devices, helping them to make more accurate diagnoses and to perform more precise image-guided interventions.

  9. X-Ray Beam Studies of Charge Sharing in Small Pixel, Spectroscopic, CdZnTe Detectors

    Science.gov (United States)

    Allwork, Christopher; Kitou, Dimitris; Chaudhuri, Sandeep; Sellin, Paul J.; Seller, Paul; Veale, Matthew C.; Tartoni, Nicola; Veeramani, Perumal

    2012-08-01

    Recent advances in the growth of CdZnTe material have allowed the development of small pixel, spectroscopic, X-ray imaging detectors. These detectors have applications in a diverse range of fields such as medical, security and industrial sectors. As the size of the pixels decreases relative to the detector thickness, the probability that charge is shared between multiple pixels increases due to the non zero width of the charge clouds drifting through the detector. These charge sharing events will result in a degradation of the spectroscopic performance of detectors and must be considered when analyzing the detector response. In this paper charge sharing and charge loss in a 250 μm pitch CdZnTe pixel detector has been investigated using a mono-chromatic X-ray beam at the Diamond Light Source, U.K. Using a 20 μm beam diameter the detector response has been mapped for X-ray energies both above (40 keV) and below (26 keV) the material K-shell absorption energies to study charge sharing and the role of fluorescence X-rays in these events.

  10. Influence of backscattering on the spatial resolution of semiconductor X-ray detectors

    International Nuclear Information System (INIS)

    Hoheisel, M.; Korn, A.; Giersch, J.

    2005-01-01

    Pixelated X-ray detectors using semiconductor layers or scintillators as absorbers are widely used in high-energy physics, medical diagnosis, or non-destructive testing. Their good spatial resolution performance makes them particularly suitable for applications where fine details have to be resolved. Intrinsic limitations of the spatial resolution have been studied in previous simulations. These simulations focused on interactions inside the conversion layer. Transmitted photons were treated as a loss. In this work, we also implemented the structure behind the conversion layer to investigate the impact of backscattering inside the detector setup. We performed Monte Carlo simulations with the program ROSI (Roentgen Simulation) which is based on the well-established EGS4 algorithm. Line-spread functions of different fully implemented detectors were simulated. In order to characterize the detectors' spatial resolution, the modulation transfer functions (MTF) were calculated. The additional broadening of the line-spread function by carrier transport has been ignored in this work. We investigated two different detector types: a directly absorbing pixel detector where a semiconductor slab is bump-bonded to a readout ASIC such as the Medipix-2 setup with Si or GaAs as an absorbing semiconductor layer, and flat-panel detectors with a Se or a CsI converter. We found a significant degradation of the MTF compared to the case without backscattering. At energies above the K-edge of the backscattering material the spatial resolution drops and can account for the observed low-frequency drop of the MTF. Ignoring this backscatter effect might lead to misinterpretations of the charge sharing effect in counting pixel detectors

  11. Development of a fast pixel array detector for use in microsecond time-resolved x-ray diffraction

    International Nuclear Information System (INIS)

    Barna, S.L.; Gruner, S.M.; Shepherd, J.A.

    1995-01-01

    A large-area pixel x-ray detector is being developed to collect eight successive frames of wide dynamic range two-dimensional images at 200kHz rates. Such a detector, in conjunction with a synchrotron radiation x-ray source, will enable time-resolved x-ray studies of proteins and other materials on time scales which have previously been inaccessible. The detector will consist of an array of fully-depleted 150 micron square diodes connected to a CMOS integrated electronics layer with solder bump-bonding. During each framing period, the current resulting from the x-rays stopped in the diodes is integrated in the electronics layer, and then stored in one of eight storage capacitors underneath the pixel. After the last frame, the capacitors are read out at standard data transmission rates. The detector has been designed for a well-depth of at least 10,000 x-rays (at 20keV), and a noise level of one x-ray. Ultimately, the authors intend to construct a detector with over one million pixels (1024 by 1024). They present the results of their development effort and various features of the design. The electronics design is discussed, with special attention to the performance requirements. The choice and design of the detective diodes, as they relate to x-ray stopping power and charge collection, are presented. An analysis of various methods of bump bonding is also presented. Finally, the authors discuss the possible need for a radiation-blocking layer, to be placed between the electronics and the detective layer, and various methods they have pursued in the construction of such a layer

  12. Modeling quantum noise of phosphors used in medical X-ray imaging detectors

    CERN Document Server

    Kalivas, N; Cavouras, D; Costaridou, L; Nomicos, C D; Panayiotakis, G S

    1999-01-01

    The noise properties of the granular phosphor screens, which are utilized in X-ray imaging detectors, are studied in terms of the quantum noise transfer function (QNTF). An analytical model, taking into account the effect of K-characteristic X-rays reabsorption within the phosphor material and the optical properties of the phosphor, was developed. The optical properties of the phosphor material required by the model were obtained from literature, except for the optical diffusion length (sigma) that was determined by data fitting and was found to be 26 cm sup 2 /g. The deviation between theoretical and experimental data is sigma depended. Specifically for sigma=26 cm sup 2 /g and sigma=25 cm sup 2 /g the respective deviations between experimental and predicted results were 0.698% and -1.597%. However for relative differences in sigma more than 15% from the value 26 cm sup 2 /g, the corresponding deviations exceed by 6 times the value of 0.698%. The model was tested via comparison to experimental results obtain...

  13. Application of the alanine detector to gamma-ray, X-ray and fast neutron dosimetry

    International Nuclear Information System (INIS)

    Waligorski, M.P.R.; Hansen, J.W.; Byrski, E.

    1987-01-01

    A dosimeter based on alanine has been developed at the INP in Krakow and at Risoe National Laboratory. Due to its near tissue-equivalence and stability of signal, measured using ESR spectrometry at room temperature, this free-radical amino-acid dosimetric system is particularly suitable for measuring X-ray, gamma-ray and fast neutron doses in the range 10-10 5 Gy. The relative effectiveness (with respect to 60 Co γ-rays) of the alanine dosimeter to 250 kVp X-rays and to cyclotron-produced fast neutrons (mean neutron energy 5.6 MeV) is measured to be 0.76± 0.06 and 0.60±0.05, respectively. The suitability of the alanine dosimeter for intercomparison gamma-ray dosimetry is also shown. The estimated absolute difference between 60 Co dosimetry at Risoe National Laboratory and at the Centre of Oncology in Krakow is about 5%, somewhat more than the experimental uncertainty. These results are based on ESR measurements performed in Krakow on about 25% of the exposed detectors. 28 refs., 2 figs., 3 tabs. (author)

  14. A novel solution to the gated x-ray detector gain droop problem

    International Nuclear Information System (INIS)

    Oertel, J. A.; Archuleta, T. N.

    2014-01-01

    Microchannel plate (MCP), microstrip transmission line based, gated x-ray detectors used at the premier ICF laser facilities have a drop in gain as a function of mircostrip length that can be greater than 50% over 40 mm. These losses are due to ohmic losses in a microstrip coating that is less than the optimum electrical skin depth. The electrical skin depth for a copper transmission line at 3 GHz is 1.2 μm while the standard microstrip coating thickness is roughly half a single skin depth. Simply increasing the copper coating thickness would begin filling the MCP pores and limit the number of secondary electrons created in the MCP. The current coating thickness represents a compromise between gain and ohmic loss. We suggest a novel solution to the loss problem by overcoating the copper transmission line with five electrical skin depths (∼6 μm) of Beryllium. Beryllium is reasonably transparent to x-rays above 800 eV and would improve the carrier current on the transmission line. The net result should be an optically flat photocathode response with almost no measurable loss in voltage along the transmission line

  15. High resolution X-ray detector for synchrotron-based microtomography

    CERN Document Server

    Stampanoni, M; Wyss, P; Abela, R; Patterson, B; Hunt, S; Vermeulen, D; Rueegsegger, P

    2002-01-01

    Synchrotron-based microtomographic devices are powerful, non-destructive, high-resolution research tools. Highly brilliant and coherent X-rays extend the traditional absorption imaging techniques and enable edge-enhanced and phase-sensitive measurements. At the Materials Science Beamline MS of the Swiss Light Source (SLS), the X-ray microtomographic device is now operative. A high performance detector based on a scintillating screen optically coupled to a CCD camera has been developed and tested. Different configurations are available, covering a field of view ranging from 715x715 mu m sup 2 to 7.15x7.15 mm sup 2 with magnifications from 4x to 40x. With the highest magnification 480 lp/mm had been achieved at 10% modulation transfer function which corresponds to a spatial resolution of 1.04 mu m. A low-noise fast-readout CCD camera transfers 2048x2048 pixels within 100-250 ms at a dynamic range of 12-14 bit to the file server. A user-friendly graphical interface gives access to the main parameters needed for ...

  16. Mapping the large area straw detectors of the COMPASS experiment with X-rays

    CERN Document Server

    Platzer, Klaus; Dünnweber, Wolfgang; Faessler, Martin A; Geyer, Reiner; Ilgner, C; Peshekhonov, Vladimir D; Wellenstein, Hermann

    2005-01-01

    In the COMPASS experiment at CERN, large straw drift tube detectors are used for large-angle tracking. To minimize the total areal density, a self supporting structure of thin-walled plastic tubes was chosen and, hence, a loss in mechanical precision was accepted. A complete mapping of the anode wire coordinate grid was required. An X-ray apparatus using a charge-coupled device (CCD) as imaging detector was built to investigate the mechanical properties and to calibrate (offline) the wire positions. Deviations of typically 200-400 mu m from the nominal positions, defined by equal spacing, are found across the detector area of 8 m/sup 2/. With a calibration method based on high-resolution CCD imaging and pattern recognition algorithms, the absolute wire coordinates are determined with an accuracy better than 30 mu m across the whole detector area. Temperature effects are clearly seen. Their inhomogenity limits the achievable accuracy to about 50 mu m under realistic experimental conditions, which is sufficient...

  17. A readout system for position sensitive measurements of X-ray using silicon strip detectors

    CERN Document Server

    Dabrowski, W; Grybos, P; Idzik, M; Kudlaty, J

    2000-01-01

    In this paper we describe the development of a readout system for X-ray measurements using silicon strip detectors. The limitation concerning the inherent spatial resolution of silicon strip detectors has been evaluated by Monte Carlo simulation and the results are discussed. The developed readout system is based on the binary readout architecture and consists of two ASICs: RX32 front-end chip comprising 32 channels of preamplifiers, shapers and discriminators, and COUNT32 counter chip comprising 32 20-bit asynchronous counters and the readout logic. This work focuses on the design and performance of the front-end chip. The RX32 chip has been optimised for a low detector capacitance, in the range of 1-3 pF, and high counting rate applications. It can be used with DC coupled detectors allowing the leakage current up to a few nA per strip. For the prototype chip manufactured in a CMOS process all basic parameters have been evaluated by electronic measurements. The noise below 140 el rms has been achieved for a ...

  18. Electrically-cooled HPGe detector for advanced x-ray spectroscopy and imaging

    Energy Technology Data Exchange (ETDEWEB)

    Marian, V.; Clauss, J.; Pirard, B.; Quirin, P.; Flamanc, J.; Lampert, M.O. [CANBERRA France, Parc des Tanneries, 1, chemin de la roseraie, 67380 Lingolsheim (France)

    2015-07-01

    High Purity Germanium (HPGe) detectors are used for high-resolution x- and gamma-ray spectroscopy. For their operation, the necessary cryogenic cooling is performed with liquid nitrogen or with electromechanical coolers. Although mature and industrialized solutions, most of HPGe detectors integrating electrical coolers present a limited spectroscopic performance due to the generated mechanical vibration and electromagnetic interference. This paper describes a novel HPGe detector, specifically designed to address the challenges of ultimate x-ray spectroscopy and imaging applications. Due to the stringent demands associated with nano-scale imaging in synchrotron applications, a custom-designed cryostat was built around a Canberra CP5-Plus electrical cooler featuring extremely low vibration levels and high cooling power. The heat generated by the cryo-cooler itself, as well as the electronics, is evacuated via an original liquid cooling circuit. This architecture can also be used to address high ambient temperature, which does not allow conventional cryo-coolers to work properly. The multichannel detector head can consist of a segmented monolithic HPGe sensor, or several closely packed sensors. Each sensor channel is read out by state-of-the-art pulse-reset preamplifiers in order to achieve excellent energy resolution for count rates in excess of 1 Mcps. The sensitive electronics are located in EMI-proof housings to avoid any interference from other devices on a beam-line. The front-end of the detector is built using selected high-purity materials and alloys to avoid any fluorescence effects. We present a detailed description of the detector design and we report on its performance. A discussion is also given on the use of electrically cooled HPGe detectors for applications requiring ultimate energy resolution, such as synchrotron, medicine or nuclear industry. (authors)

  19. Asymmetry of characteristic X-ray peaks obtained by a Si(Li) detector

    Energy Technology Data Exchange (ETDEWEB)

    Visnovezky, Claudia [Facultad de Matematica, Astronomia y Fisica, Universidad Nacional de Cordoba, Ciudad Universitaria, 5000, Cordoba (Argentina)], E-mail: cavy3@hotmail.com; Limandri, Silvina [Facultad de Matematica, Astronomia y Fisica, Universidad Nacional de Cordoba, Ciudad Universitaria, 5000, Cordoba (Argentina)], E-mail: silvilimandri@hotmail.com; Canafoglia, Maria Elena [Centro de Investigacion y Desarrollo en Ciencias Aplicadas Dr. Jorge Ronco, Calle 47 No 257, 1900 La Plata, Argentina, Facultad de Ciencias Exactas y Facultad de Ingenieria de la UNLP, La Plata (Argentina); Bonetto, Rita [Centro de Investigacion y Desarrollo en Ciencias Aplicadas Dr. Jorge Ronco, Calle 47 No 257, 1900 La Plata, Argentina, Facultad de Ciencias Exactas y Facultad de Ingenieria de la UNLP, La Plata (Argentina); Consejo Nacional de Investigaciones Cientificas y Tecnicas de la Republica Argentina (Argentina)], E-mail: bonetto@quimica.unlp.edu.ar; Trincavelli, Jorge [Facultad de Matematica, Astronomia y Fisica, Universidad Nacional de Cordoba, Ciudad Universitaria, 5000, Cordoba (Argentina); Consejo Nacional de Investigaciones Cientificas y Tecnicas de la Republica Argentina (Argentina)], E-mail: jorge@quechua.fis.uncor.edu

    2007-05-15

    The asymmetry of the characteristic X-ray peaks obtained using a Si(Li) detector is mainly due to incomplete charge collection. Impurities and defects in the crystalline structure of Si can act as 'traps' for holes and electrons in their trip toward the detector electrodes. Therefore, the collected charge, and consequently the detected energy, is smaller than the expected one. The global effect is that peaks may present a 'tail' toward the low energy side. This phenomenon is more important for low energies (lower than 2.3 keV, in the case of the detector characterized). In this work, the parameters related to peak asymmetry were studied, allowing a better understanding of the trapping process mentioned above. For this purpose, spectra from mono- and multi-element samples were collected for elements with atomic number between 7 and 20. In order to describe the shape of the characteristic K peaks as a function of its energy, an asymmetric correction to a Gaussian function was proposed. Spectra were obtained by electron probe microanalysis for incidence energies between 5 and 25 keV using an energy dispersive spectrometer equipped with an ultra-thin window Si(Li) detector. It was observed that the area corresponding to the asymmetric correction exhibits an energy dependence similar to that of the mass absorption coefficient of the detector material. In addition, other two spectrometers were used to investigate the dependence of tailing on the detection system. When two spectrometers with the same kind of detector and different pulse processors were compared, peaks were more asymmetric for lower peaking time values. When two different detectors were used, differences were even more important.

  20. High Energy Replicated Optics to Explore the Sun: Hard X-ray balloon-borne telescope

    Science.gov (United States)

    Gaskin, J.; Apple, J.; Chavis, K. S.; Dietz, K.; Holt, M.; Koehler, H.; Lis, T.; O'Connor, B.; Otero, M. R.; Pryor, J.; Ramsey, B.; Rinehart-Dawson, M.; Smith, L.; Sobey, A.; Wilson-Hodge, C.; Christe, S.; Cramer, A.; Edgerton, M.; Rodriguez, M.; Shih, A.; Gregory, D.; Jasper, J.; Bohon, S.

    Set to fly in the Fall of 2013 from Ft. Sumner, NM, the High Energy Replicated Optics to Explore the Sun (HEROES) mission is a collaborative effort between the NASA Marshall Space Flight Center and the Goddard Space Flight Center to upgrade an existing payload, the High Energy Replicated Optics (HERO) balloon-borne telescope, to make unique scientific measurements of the Sun and astrophysical targets during the same flight. The HEROES science payload consists of 8 mirror modules, housing a total of 109 grazing-incidence optics. These modules are mounted on a carbon-fiber - and Aluminum optical bench 6 m from a matching array of high pressure xenon gas scintillation proportional counters, which serve as the focal-plane detectors. The HERO gondola utilizes a differential GPS system (backed by a magnetometer) for coarse pointing in the azimuth and a shaft angle encoder plus inclinometer provides the coarse elevation. The HEROES payload will incorporate a new solar aspect system to supplement the existing star camera, for fine pointing during both the day and night. A mechanical shutter will be added to the star camera to protect it during solar observations. HEROES will also implement two novel alignment monitoring system that will measure the alignment between the optical bench and the star camera and between the optics and detectors for improved pointing and post-flight data reconstruction. The overall payload will also be discussed. This mission is funded by the NASA HOPE (Hands On Project Experience) Training Opportunity awarded by the NASA Academy of Program/Project and Engineering Leadership, in partnership with NASA's Science Mission Directorate, Office of the Chief Engineer and Office of the Chief Technologist.

  1. Characterization of a spectroscopic detector for application in x-ray computed tomography

    Science.gov (United States)

    Dooraghi, Alex A.; Fix, Brian J.; Smith, Jerel A.; Brown, William D.; Azevedo, Stephen G.; Martz, Harry E.

    2017-09-01

    Recent advances in cadmium telluride (CdTe) energy-discriminating pixelated detectors have enabled the possibility of Multi-Spectral X-ray Computed Tomography (MSXCT) to incorporate spectroscopic information into CT. MultiX ME 100 V2 is a CdTe-based spectroscopic x-ray detector array capable of recording energies from 20 to 160 keV in 1.1 keV energy bin increments. Hardware and software have been designed to perform radiographic and computed tomography tasks with this spectroscopic detector. Energy calibration is examined using the end-point energy of a bremsstrahlung spectrum and radioisotope spectral lines. When measuring the spectrum from Am-241 across 500 detector elements, the standard deviation of the peak-location and FWHM measurements are +/- 0.4 and +/- 0.6 keV, respectively. As these values are within the energy bin size (1.1 keV), detector elements are consistent with each other. The count rate is characterized, using a nonparalyzable model with a dead time of 64 +/- 5 ns. This is consistent with the manufacturer's quoted per detector-element linear-deviation at 2 Mpps (million photons per sec) of 8.9 % (typical) and 12 % (max). When comparing measured and simulated spectra, a low-energy tail is visible in the measured data due to the spectral response of the detector. If no valid photon detections are expected in the low-energy tail, then a background subtraction may be applied to allow for a possible first-order correction. If photons are expected in the low-energy tail, a detailed model must be implemented. A radiograph of an aluminum step wedge with a maximum height of 20 mm shows an underestimation of attenuation by about 10 % at 60 keV. This error is due to partial energy deposition from higher energy (>60 keV) photons into a lower-energy ( 60 keV) bin, reducing the apparent attenuation. A radiograph of a polytetrafluoroethylene (PTFE) cylinder taken using a bremsstrahlung spectrum from an x-ray voltage of 100 kV filtered by 1.3 mm Cu is

  2. Characterization of a spectroscopic detector for application in x-ray computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, A. A. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Fix, B. J. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Smith, J. A. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Brown, W. D. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Azevedo, S. G. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Martz, H. E. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2017-04-21

    Recent advances in cadmium telluride (CdTe) energy-discriminating pixelated detectors have enabled the possibility of Multi-Spectral X-ray Computed Tomography (MSXCT) to incorporate spectroscopic information into CT. MultiX ME 100 V2 is a CdTe-based spectroscopic x-ray detector array capable of recording energies from 20 to 160 keV in 1.1 keV energy bin increments. Hardware and software have been designed to perform radiographic and computed tomography tasks with this spectroscopic detector. Energy calibration is examined using the end-point energy of a bremsstrahlung spectrum and radioisotope spectral lines. When measuring the spectrum from Am-241 across 500 detector elements, the standard deviation of the peak-location and FWHM measurements are ±0.4 and ±0.6 keV, respectively. As these values are within the energy bin size (1.1 keV), detector elements are consistent with each other. The count rate is characterized, using a nonparalyzable model with a dead time of 64 ± 5 ns. This is consistent with the manufacturer’s quoted per detector-element linear-deviation at 2 Mpps (million photons per sec) of 8.9% (typical) and 12% (max). When comparing measured and simulated spectra, a low-energy tail is visible in the measured data due to the spectral response of the detector. If no valid photon detections are expected in the low-energy tail, then a background subtraction may be applied to allow for a possible first-order correction. If photons are expected in the low-energy tail, a detailed model must be implemented. A radiograph of an aluminum step wedge with a maximum height of about 20 mm shows an underestimation of attenuation by about 10% at 60 keV. This error is due to partial energy deposition from higher-energy (> 60 keV) photons into a lower-energy (~60 keV) bin, reducing the apparent attenuation. A radiograph of a PTFE cylinder taken using a bremsstrahlung spectrum from an x-ray voltage of 100 kV filtered by 1.3 mm Cu is reconstructed using Abel inversion

  3. High-Resolution Detector for At-Wavelength Metrology of X-Ray Optics, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — Since the launch of the first X-ray focusing telescope in 1963, the development of grazing incidence X-ray optics has been crucial to the development of the field of...

  4. High-Resolution Detector for At-Wavelength Metrology of X-Ray Optics, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Since the launch of the first X-ray focusing telescope in 1963, the development of grazing incidence X-ray optics has been crucial to the development of the field of...

  5. Urgent X-ray examination of new-born babies. 1. Oesophagus-stomach

    Energy Technology Data Exchange (ETDEWEB)

    Ponhold, W; Poplavski, K [Vienna Univ. (Austria). Kinderklinik

    1981-01-01

    This paper provides the paediatrician using radiology and his assistants with exact and easily-understood instructions as to how to carry out an urgent X-ray examination of the oesophagus and the stomach. As far as an X-ray examination for the diagnosis of an atresia of the oesophagus, an oesophago-tracheo fistula and a pylorusstenosis, are concerned, fluoroscopy is in the first instance unnecessary, provided that an exact sequence of steps is kept to. Only in cases of long-term check-ups after oesophagus operations and oesophago-tracheo fistulas that are difficult to prove fluoroscopy is necessary. By carrying out the examinations suggested here, a correct diagnosis can be reached soon and exposure of the child to radiations kept to a minimum.

  6. Research of x-ray nondestructive detector for high-speed running conveyor belt with steel wire ropes

    Science.gov (United States)

    Wang, Junfeng; Miao, Changyun; Wang, Wei; Lu, Xiaocui

    2008-03-01

    An X-ray nondestructive detector for high-speed running conveyor belt with steel wire ropes is researched in the paper. The principle of X-ray nondestructive testing (NDT) is analyzed, the general scheme of the X-ray nondestructive testing system is proposed, and the nondestructive detector for high-speed running conveyor belt with steel wire ropes is developed. The hardware of system is designed with Xilinx's VIRTEX-4 FPGA that embeds PowerPC and MAC IP core, and its network communication software based on TCP/IP protocol is programmed by loading LwIP to PowerPC. The nondestructive testing of high-speed conveyor belt with steel wire ropes and network transfer function are implemented. It is a strong real-time system with rapid scanning speed, high reliability and remotely nondestructive testing function. The nondestructive detector can be applied to the detection of product line in industry.

  7. Real time 2 dimensional detector for charged particle and soft X-ray images

    International Nuclear Information System (INIS)

    Ishikawa, M.; Ito, M.; Endo, T.; Oba, K.

    1995-01-01

    The conventional instruments used in experiments for the soft X-ray region such as X-ray diffraction analysis are X-ray films or imaging plates. However, these instruments are not suitable for real time observation. In this paper, newly developed imaging devices will be presented, which have the capability to take X-ray images in real time with a high detection efficiency. Also, another capability, to take elementary particle tracking images, is described. (orig.)

  8. Active Detectors for Plasma Soft X-Ray Detection at PALS

    Directory of Open Access Journals (Sweden)

    C. Granja

    2010-01-01

    Full Text Available This paper summarizes the work carried out for an experimental study of low-energy nuclear excitation by laser-produced plasma at the PALS Prague laser facility. We describe the adaptation and shielding of single-quantum active radiation detectors developed at IEAP CTU Prague to facilitate their operation inside the laser interaction chamber in the vicinity of the plasma target. The goal of this effort is direct real-time single-quantum detection of plasma soft X-ray radiation with energy above a few keV and subsequent identification of the decay of the excited nuclear states via low-energy gamma rays in a highly radiative environment with strong electromagnetic interference.

  9. Superficial x-ray in-vivo dosimetry with MOSFET detectors

    International Nuclear Information System (INIS)

    Cheung, T.; Yu, P.K.N.; Butson, M.J.; Cancer Services, Wollongong, NSW

    2004-01-01

    Full text: This note investigates in-vivo dosimetry using a Metal Oxide Semiconductor Field Effect Transistor (MOSFET) for radiotherapy treatment at superficial and orthovoltage x-ray energies. This was performed within one fraction of the patient's treatment. Standard measurements along with energy response of the detector are given. Results showed that the MOSFET measurements in-vivo agreed with calculated results on average within ± 5.6% over all superficial and orthovoltage energies. These variations were slightly larger than TLD results with variations between measured and calculated results being ± 5.0% for the same patient measurements. The MOSFET device provides adequate in-vivo dosimetry for superficial and orthovoltage energy treatments with the accuracy of the measurements seeming to be relatively on par with TLD in our case. The MOSFET does have the advantage of returning a relatively immediate dosimetric result after irradiation. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine

  10. Time response of fast-gated microchannel plates used as x-ray detectors

    International Nuclear Information System (INIS)

    Turner, R.E.; Bell, P.; Hanks, R.; Kilkenny, J.D.; Landen, N.; Power, G.; Wiedwald, J.; Meier, M.

    1990-01-01

    We report measurements of the time response of fast-gated, micro- channel plate (MCP) detectors, using a <10 ps pulsewidth ultra-violet laser and an electronic sampling system to measure time resolutions to better than 25 ps. The results show that framing times of less than 100 ps are attainable with high gain. The data is compared to a Monte Carlo calculation, which shows good agreement. We also measured the relative sensitivity as a function of DC bias, and saturation effects for large signal inputs. In part B, we briefly describe an electrical ''time-of-flight'' technique, which we have used to measure the response time of a fast-gated microchannel plate (MCP). Thinner MCP's than previously used have been tested, and, as expected, show fast gating times and smaller electron multiplication. A preliminary design for an x-ray pinhole camera, using a thin MCP, is presented. 7 refs., 6 figs

  11. Substrate and Passivation Techniques for Flexible Amorphous Silicon-Based X-ray Detectors.

    Science.gov (United States)

    Marrs, Michael A; Raupp, Gregory B

    2016-07-26

    Flexible active matrix display technology has been adapted to create new flexible photo-sensing electronic devices, including flexible X-ray detectors. Monolithic integration of amorphous silicon (a-Si) PIN photodiodes on a flexible substrate poses significant challenges associated with the intrinsic film stress of amorphous silicon. This paper examines how altering device structuring and diode passivation layers can greatly improve the electrical performance and the mechanical reliability of the device, thereby eliminating one of the major weaknesses of a-Si PIN diodes in comparison to alternative photodetector technology, such as organic bulk heterojunction photodiodes and amorphous selenium. A dark current of 0.5 pA/mm² and photodiode quantum efficiency of 74% are possible with a pixelated diode structure with a silicon nitride/SU-8 bilayer passivation structure on a 20 µm-thick polyimide substrate.

  12. Radiation effects on the proportional counter X-ray detectors on board the NEAR spacecraft

    International Nuclear Information System (INIS)

    Floyd, S.R.; Trombka, J.I.; Leidecker, H.W.; Clark, P.E.; Starr, R.; Goldsten, J.O.; Roth, D.R.

    1999-01-01

    The X-ray proportional counters on board the Near Earth Asteroid Rendezvous (NEAR) spacecraft have exhibited a resolution degradation and recovery phenomenon several times during the long cruise phase of the mission. The resolution is checked periodically by commanding an 55 Fe source into the window area. The degradation is seen as a low energy tailing of the 5.9 keV photopeak. Two events have occurred which provided good spectral data for better understanding the degradation phenomenon. In November 1997 a large solar particle event occurred that degraded the resolution and excited copper in the collimator. Eventually the detectors returned to normal. In January 1998 the spacecraft performed an Earth swingby gravity assist maneuver. The near Earth environment excited the magnesium and aluminum in the filter elements. The copper line was also produced. The NEAR spacecraft was launched in February 1996 and will rendezvous and orbit the asteroid 433 Eros in early 1999

  13. Feasibility study of flexible flat-panel X-ray detectors for digital radiography

    International Nuclear Information System (INIS)

    Joe, Ok La; Yun, Seung Man; Kim, Ho Kyung

    2010-01-01

    Flexible flat-panel detectors (FPDs), which utilize both organic photodiode (OPD) and organic thin-film transistor (OTFT) technologies, are recently concerned in digital radiography. The flexible FPD has several potential advantages, such as high accessibility to patient, avoidance of geometrical burr due to the oblique angle incidence of X-ray, great reduction in manufacturing cost due to jet-printing. At once, The OPD/OTFT arrays were fabricated by jet-printing techniques, mechanical robustness due to plastic substrates, and so on. In this study, we have investigated the feasibility of flexible FPD by comparing theoretical detective quantum efficiency (DQE) with that of the conventional amorphous silicon-based FPD. We chose copper phthalocyanine-fullerene (CuPc-C60) organic materials for the construction of the flexible FPD. DQE was calculated by the linear-systems transfer theory

  14. Using phonon pulses to characterise superconducting tunnel junction (STJ) X-ray detectors

    International Nuclear Information System (INIS)

    Steele, A.; Kozorezov, A.G.; Boyd, P.; Wigmore, J.K.; Poelaert, A.; Peacock, A.; Hartog, R. den

    2000-01-01

    Nanosecond phonon pulse experiments have been used to determine fundamental parameters of STJs relevant to their use as X-ray photon detectors. A non-equilibrium distribution of phonons is used to generate an excess non-equilibrium quasi-particle (qp) density in the STJ base electrode. The time dependence of the subsequent current signal is given by the sum of two exponential contributions which depend solely on the qp loss rates and tunnel rates for the top and base electrode of the device. Hence, four fundamental STJ parameters can be determined from measurements of the exponential time constants and pre-exponential current amplitudes. The technique outlined here is demonstrated by data taken on a high-quality 50 μmx50 μm niobium-based STJ

  15. Application of charge coupled devices as spatially-resolved detectors for X-ray spectrograph

    Energy Technology Data Exchange (ETDEWEB)

    Attelan-Langlet, S; Etlicher, B [Ecole Polytechnique, Palaiseau (France); Mishenskij, V O; Papazyan, Yu V; Smirnov, V P; Volkov, G S; Zajtsev, V I [Inst. for Thermonuclear and Innovation Investigations, Troitsk (Russian Federation)

    1997-12-31

    An X-ray crystal spectrograph which contains a CCD linear array as the position-sensitive detector is described. Radiation detection is performed directly onto CCD. The spectrograph has a limit of sensitivity at about 2 J/(A.ster), spectral resolution about 1000 and dynamic range 100-120. The device operates on-line with IBM-PC based control system. Software provides all data acquisition and treatment. Output spectra are presented in absolute units. The device was used during composite Z-pinch experiments at pulse-power installations ``Angara-5-1`` (TRINITI, Troitsk, Russia) and ``GAEL`` (Ecole Polytechnique, Palaiseau, France). Currently the spectrograph is included in the set of diagnostics of the ``Angara-5-1`` facility. Some of the spectra obtained are presented and discussed. (author). 4 figs., 9 refs.

  16. Time-resolved and position-resolved X-ray spectrometry with a pixelated detector

    Energy Technology Data Exchange (ETDEWEB)

    Sievers, Peter

    2012-12-07

    The aim of the work presented here was to measure X-ray spectra with a pixelated detector. Due to effects in the sensor the spectrum cannot be measured directly and has to be calculated by a deconvolution of the measured data. In the scope of this work the deconvolution of the measured spectra could be enhanced considerably by - amongst other things - the introduction of the Bayesian deconvolution method. Those improvements opened the possibilities for further measurements. For the measurements the detectors of the Medipix family have been used. They are nowadays used for a wide range of applications and scientific research. Their main advantage is the very high position resolution gained by a pixel pitch of 55 μm and a high number of 65536 pixels. The Timepix detector has, in particular, two special possibilities of measurement: the ToA mode and the ToT mode. In ToA mode the arrival time of an impinging photon is measured and in ToT mode the amount of deposited charge is measured. The most common method of operation is counting the number of impinging photons that release a charge higher than a preset threshold in each pixel. As this released charge is proportional to the energy deposition of the impinging photon, one can perform energy-sensitive measurements. To perform the deconvolution of the measured energy distribution there is a need of an energy response matrix describing the detector response on radiation. For some detectors it is possible to obtain an analytic model of the response functions. Due to the high discrepancy between the impinging spectrum and the measured spectrum in case of detectors of the Medipix family, there is so far no analytic model. Thus, the detector response has to be simulated. As I could improve the precision of the measurement quite extensively, I also intended to tune the simulation with more accurate and appropriate models to gain the same level of accuracy. The results of measurement and simulation have then been compared and

  17. Time-resolved and position-resolved X-ray spectrometry with a pixelated detector

    International Nuclear Information System (INIS)

    Sievers, Peter

    2012-01-01

    The aim of the work presented here was to measure X-ray spectra with a pixelated detector. Due to effects in the sensor the spectrum cannot be measured directly and has to be calculated by a deconvolution of the measured data. In the scope of this work the deconvolution of the measured spectra could be enhanced considerably by - amongst other things - the introduction of the Bayesian deconvolution method. Those improvements opened the possibilities for further measurements. For the measurements the detectors of the Medipix family have been used. They are nowadays used for a wide range of applications and scientific research. Their main advantage is the very high position resolution gained by a pixel pitch of 55 μm and a high number of 65536 pixels. The Timepix detector has, in particular, two special possibilities of measurement: the ToA mode and the ToT mode. In ToA mode the arrival time of an impinging photon is measured and in ToT mode the amount of deposited charge is measured. The most common method of operation is counting the number of impinging photons that release a charge higher than a preset threshold in each pixel. As this released charge is proportional to the energy deposition of the impinging photon, one can perform energy-sensitive measurements. To perform the deconvolution of the measured energy distribution there is a need of an energy response matrix describing the detector response on radiation. For some detectors it is possible to obtain an analytic model of the response functions. Due to the high discrepancy between the impinging spectrum and the measured spectrum in case of detectors of the Medipix family, there is so far no analytic model. Thus, the detector response has to be simulated. As I could improve the precision of the measurement quite extensively, I also intended to tune the simulation with more accurate and appropriate models to gain the same level of accuracy. The results of measurement and simulation have then been compared and

  18. Pulsed X-ray radiography of a gas jet target for laser-matter interaction experiments with the use of a CCD detector

    International Nuclear Information System (INIS)

    Rakowski, R.; Bartnik, A.; Fiedorowicz, H.; Jarocki, R.; Kostecki, J.; MikoIajczyk, J.; Szczurek, A.; Szczurek, M.; Foeldes, I.B.; Toth, Zs.

    2005-01-01

    Characterization of gas jet targets has been carried out using pulsed X-ray radiography. A laser-plasma X-ray source was applied for backlighting of the targets to obtain X-ray shadowgraphs registered with a CCD detector. From the shadowgraphs, characteristics of the targets were determined

  19. Investigation of photon counting pixel detectors for X-ray spectroscopy and imaging

    Energy Technology Data Exchange (ETDEWEB)

    Talla, Patrick Takoukam

    2011-04-07

    The Medipix2 and Medipix3 detectors are hybrid pixelated photon counting detectors with a pixel pitch of 55 {mu}m. The sensor material used in this thesis was silicon. Because of their small pixel size they suffer from charge sharing i.e. an incoming photon can be registered by more than one pixel. In order to correct for charge sharing due to lateral diffusion of charge carriers, the Medipix3 detector was developed: with its Charge Summing Mode, the charge collected in a cluster of 2 x 2 pixel is added up and attributed to only one pixel whose counter is incremented. The adjustable threshold of the detectors allows to count the photons and to gain information on their energy. The main purposes of the thesis are to investigate spectral and imaging properties of pixelated photon counting detectors from the Medipix family such as Medipix2 and Medipix3. The investigations are based on simulations and measurements. In order to investigate the spectral properties of the detectors measurements were performed using fluorescence lines of materials such as molybdenum, silver but also some radioactive sources such as Am-241 or Cd-109. From the measured data, parameters like the threshold dispersion and the gain variation from pixel-to-pixel were extracted and used as input in the Monte Carlo code ROSI to model the responses of the detector to monoenergetic photons. The measured data are well described by the simulations for Medipix2 and for Medipix3 operating in Charge Summing Mode. Due to charge sharing and due to the energy dependence of attenuation processes in silicon and to Compton scattering the incoming and the measured spectrum differ substantially from each other. Since the responses to monoenergetic photons are known, a deconvolution was performed to determine the true incoming spectrum. Several direct and iterative methods were successfully applied on measured and simulated data of an X-ray tube and radioactive sources. The knowledge of the X-ray spectrum is

  20. Investigation of photon counting pixel detectors for X-ray spectroscopy and imaging

    International Nuclear Information System (INIS)

    Talla, Patrick Takoukam

    2011-01-01

    The Medipix2 and Medipix3 detectors are hybrid pixelated photon counting detectors with a pixel pitch of 55 μm. The sensor material used in this thesis was silicon. Because of their small pixel size they suffer from charge sharing i.e. an incoming photon can be registered by more than one pixel. In order to correct for charge sharing due to lateral diffusion of charge carriers, the Medipix3 detector was developed: with its Charge Summing Mode, the charge collected in a cluster of 2 x 2 pixel is added up and attributed to only one pixel whose counter is incremented. The adjustable threshold of the detectors allows to count the photons and to gain information on their energy. The main purposes of the thesis are to investigate spectral and imaging properties of pixelated photon counting detectors from the Medipix family such as Medipix2 and Medipix3. The investigations are based on simulations and measurements. In order to investigate the spectral properties of the detectors measurements were performed using fluorescence lines of materials such as molybdenum, silver but also some radioactive sources such as Am-241 or Cd-109. From the measured data, parameters like the threshold dispersion and the gain variation from pixel-to-pixel were extracted and used as input in the Monte Carlo code ROSI to model the responses of the detector to monoenergetic photons. The measured data are well described by the simulations for Medipix2 and for Medipix3 operating in Charge Summing Mode. Due to charge sharing and due to the energy dependence of attenuation processes in silicon and to Compton scattering the incoming and the measured spectrum differ substantially from each other. Since the responses to monoenergetic photons are known, a deconvolution was performed to determine the true incoming spectrum. Several direct and iterative methods were successfully applied on measured and simulated data of an X-ray tube and radioactive sources. The knowledge of the X-ray spectrum is

  1. Optimization of detector pixel size for stent visualization in x-ray fluoroscopy

    International Nuclear Information System (INIS)

    Jiang Yuhao; Wilson, David L.

    2006-01-01

    Pixel size is of great interest in the flat-panel detector design because of its potential impact on image quality. In the particular case of angiographic x-ray fluoroscopy, small pixels are required in order to adequately visualize interventional devices such as guidewires and stents which have wire diameters as small as 200 and 50 μm, respectively. We used quantitative experimental and modeling techniques to investigate the optimal pixel size for imaging stents. Image quality was evaluated by the ability of subjects to perform two tasks: detect the presence of a stent and discriminate a partially deployed stent from a fully deployed one in synthetic images. With measurements at 50, 100, 200, and 300 μm, the 100 μm pixel size gave the maximum contrast sensitivity for the detection experiment with the idealized direct detector. For an idealized indirect detector with a scintillating layer, an optimal pixel size was obtained at 200 μm pixel size. A channelized human observer model predicted a peak at 150 and 170 μm, for the idealized direct and indirect detectors, respectively. With regard to the stent deployment task for both detector types, smaller pixel sizes are favored and there is a steep drop in performance with larger pixels. In general, with the increasing exposures, the model and measurements give the enhanced contrast sensitivities and a smaller optimal pixel size. The effects of electronic noise and fill factor were investigated using the model. We believe that the experimental results and human observer model predications can help guide the flat-panel detector design. In addition, the human observer model should work on the similar images and be applicable to the future model and actual flat-panel implementations

  2. Characteristics of an intrinsic germanium detector for measurement of soft x-rays from high-temperature plasmas

    International Nuclear Information System (INIS)

    Kumagai, Katsuaki; Matoba, Tohru; Funahashi, Akimasa; Kawakami, Tomohide

    1976-09-01

    An intrinsic germanium (Ge(I)) detector has been prepared for measurement of soft X-ray spectra from high-temperature tokamak plasmas. Its characteristics of photo-peak efficiency, escape-peak and Compton scattering were calibrated with standard radioisotopes and soft X-rays from the JFT-2a plasma, and compared with those of a lithium-drifted silicon (Si(Li)) detector. Features of the Ge(I) detector are as follows: (i) high detection efficiency in the high energy range, (ii) wide energy range for measurement of soft X-ray spectra, and (iii) low Compton scattering effect in measurement of continuous spectra. Its dead-layer depth is about 0.06μm, and the minimum detectable energies in the Ge(I) detector are similar to those in the Si(Li) detector. The Ge(I) detector is effective for measuring soft X-ray spectra from high-temperature tokamak plasmas. (auth.)

  3. Multielement X-ray row detector on GaAs with spatial resolution of 108 μm

    International Nuclear Information System (INIS)

    Dvoryankin, V.F.; Dikaev, Yu.M.; Krikunov, A.I.; Panova, T.M.; Telegin, A.A.

    2004-01-01

    The multielement X-ray row detector with pitch of 108 μm was made on epitaxial GaAs (p + -n-n'-n + ) structures by isotropic etching in solution HCl-KBrO 3 -H 2 O. Separation of signals from the near-by detectors is achieved by built-in guard ring on each pixel. The spatial response of the detectors was evaluated

  4. Multielement X-ray row detector on GaAs with spatial resolution of 108 {mu}m

    Energy Technology Data Exchange (ETDEWEB)

    Dvoryankin, V.F.; Dikaev, Yu.M. E-mail: ymd289@ire216.msk.ru; Krikunov, A.I.; Panova, T.M.; Telegin, A.A

    2004-09-21

    The multielement X-ray row detector with pitch of 108 {mu}m was made on epitaxial GaAs (p{sup +}-n-n'-n{sup +}) structures by isotropic etching in solution HCl-KBrO{sub 3}-H{sub 2}O. Separation of signals from the near-by detectors is achieved by built-in guard ring on each pixel. The spatial response of the detectors was evaluated.

  5. Technical characterization of five x-ray detectors for paediatric radiography applications

    Science.gov (United States)

    Marshall, N. W.; Smet, M.; Hofmans, M.; Pauwels, H.; De Clercq, T.; Bosmans, H.

    2017-12-01

    Physical image quality of five x-ray detectors used in the paediatric imaging department is characterized with the aim of establishing the range/scope of imaging performance provided by these detectors for neonatal imaging. Two computed radiography (CR) detectors (MD4.0 powder imaging plate (PIP) and HD5.0 needle imaging plate (NIP), Agfa HealthCare NV, B-2640 Mortsel, Belgium) and three flat panel detectors (FPD) (the Agfa DX-D35C and DX-D45C and the DRX-2530C (Carestream Health Inc., Rochester, NY 14608, USA)) were assessed. Physical image quality was characterized using the detector metrics given by the International Electrotechnical Commission (IEC 62220-1) to measure modulation transfer function (MTF), the noise power spectrum (NPS) and the detective quantum efficiency (DQE) using the IEC-specified beam qualities of RQA3 and RQA5. The DQE was evaluated at the normal operating detector air kerma (DAK) level, defined at 2.5 µGy for all detectors, and at factors of 1/3.2 and 3.2 times the normal level. MTF curves for the different detectors were similar at both RQA3 and RQA5 energies; the average spatial frequency for the 50% point (MTF0.5) at RQA3 was 1.26 mm-1, with a range from 1.20 mm-1 to 1.37 mm-1. The DQE of the NIP CR compared to the PIP CR was notably greater and similar to that for the FPD devices. At RQA3, average DQE for the FPD and NIP (at 0.5 mm-1 2.5 µGy) was 0.57 compared to 0.26 for the PIP CR. At the RQA5 energy, the DRX-2530C and the DX-D45C had the highest DQE (~0.6 at 0.5 mm-1 2.5 µGy). Noise separation analysis using the polynomial model showed higher electronic noise for the DX-D35C and DRX-2530C detectors; this explains the reduced DQE seen at 0.7 µGy/image. The NIP CR detector offers notably improved DQE performance compared to the PIP CR system and a value similar to the DQE for FPD devices at the RQA3 energy.

  6. Field demonstration of a portable, X-ray, K-edge heavy-metal detector

    International Nuclear Information System (INIS)

    Jensen, T.; Aljundi, T.; Whitmore, C.; Zhong, H.; Gray, J.N.

    1997-01-01

    Under the Characterization, Monitoring, and Sensor Technology Crosscutting Program, the authors have designed and built a K-edge heavy metal detector that measures the level of heavy metal contamination inside closed containers in a nondestructive, non-invasive way. The device employs a volumetric technique that takes advantage of the X-ray absorption characteristics of heavy elements, and is most suitable for characterization of contamination inside pipes, processing equipment, closed containers, and soil samples. The K-edge detector is a fast, efficient, and cost-effective in situ characterization tool. More importantly, this device will enhance personnel safety while characterizing radioactive and toxic waste. The prototype K-edge system was operated at the Materials and Chemistry Laboratory User Facility at the Oak Ridge K-25 Site during February 1997. Uranium contaminated pipes and valves from a UF 6 feed facility were inspected using the K-edge technique as well as a baseline nondestructive assay method. Operation of the K-edge detector was demonstrated for uranium contamination ranging from 10 to 6,000 mg/cm 2 and results from the K-edge measurements were found to agree very well with nondestructive assay measurements

  7. X-ray and gamma-ray standards for detector calibration

    International Nuclear Information System (INIS)

    1991-09-01

    The IAEA established a Co-ordinated Research Programme (CRP) on the Measurements and Evaluation of X- and Gamma-Ray Standards for Detector Efficiency Calibration in 1986 with the aim of alleviating the generation of such discrepancies. Within the framework of this CRP, representatives of nine research groups from six Member States and one international organization performed a number of precise measurements and systematic in-depth evaluations of the required decay data. They have also contributed to the development of evaluation methodology and measurement techniques, and stimulated a number of such studies at laboratories not directly involved in the IAEA project. The results of the work of the CRP, which was finished in 1990, are presented in this report. Recommended values of half-lives and photon emission probabilities are given for a carefully selected set of radionuclides that are suitable for detector efficiency calibration (X-rays from 5 to 90 keV and gamma-rays from 30 to about 3000 keV). Detector efficiency calibration for higher gamma-ray energies (up to 14 MeV) is also considered. The evaluation procedures used to obtain the recommended values and their estimated uncertainties are reported, and a summary of the remaining discrepancies is given. Refs and tabs

  8. Applications of pixellated GaAs X-ray detectors in a synchrotron radiation beam

    CERN Document Server

    Watt, J; Campbell, M; Mathieson, K; Mikulec, B; O'Shea, V; Passmore, M S; Schwarz, C; Smith, K M; Whitehill, C

    2001-01-01

    Hybrid semiconductor pixel detectors are being investigated as imaging devices for radiography and synchrotron radiation beam applications. Based on previous work in the CERN RD19 and the UK IMPACT collaborations, a photon counting GaAs pixel detector (PCD) has been used in an X-ray powder diffraction experiment. The device consists of a 200 mu m thick SI-LEC GaAs detector patterned in a 64*64 array of 170 mu m pitch square pixels, bump-bonded to readout electronics operating in single photon counting mode. Intensity peaks in the powder diffraction pattern of KNbO/sub 3/ have been resolved and compared with results using the standard scintillator, and a PCD predecessor (the Omega 3). The PCD shows improved speed, dynamic range, 2-D information and comparable spatial resolution to the standard scintillator based systems. It also overcomes the severe dead time limitations of the Omega 3 by using a shutter based acquisition mode. A brief demonstration of the possibilities of the system for dental radiography and...

  9. The hard X-ray response of epitaxial GaAs detectors

    CERN Document Server

    Owens, A; Kraft, S; Peacock, A; Nenonen, S; Andersson, H

    2000-01-01

    We report on hard X-ray measurements with two epitaxial GaAs detectors of active areas 2.22 mm sup 2 and thicknesses 40 and 400 mu m at the ESRF and HASYLAB synchrotron research facilities. The detectors were fabricated using high-purity material and in spite of an order of magnitude difference in depletion depths, they were found to have comparable performances with energy resolutions at -45 deg. C of approx 1 keV fwhm at 7 keV rising to approx 2 keV fwhm at 200 keV and noise floors in the range 1-1.5 keV. At energies <30 keV, the energy resolution was dominated by leakage current and electromagnetic pick-up, while at the highest energies measured, the resolutions approach the expected Fano limit (e.g., approx 1 keV near 200 keV). Both detectors are remarkably linear, with average rms non-linearities of 0.2% over the energy range 10-60 keV, which, taken in conjunction with Monte-Carlo results indicate that charge collection efficiencies must be in excess of 98%. This is consistent with material science me...

  10. Search for chameleons with an InGrid based X-ray detector at the CAST experiment

    Energy Technology Data Exchange (ETDEWEB)

    Desch, Klaus; Kaminski, Jochen; Krieger, Christoph; Schmidt, Sebastian [Physikalisches Institut, Universitaet Bonn, Nussallee 12, 53115 Bonn (Germany)

    2016-07-01

    The CERN Axion Solar Telescope (CAST) searches for axions and also other exotic particles emerging from the Sun. Chameleons, for example, are part of Dark Energy theories. Like Axions they can be converted into soft X-ray photons in a high magnetic field and should result in an X-ray spectrum peaking below 1 keV. Because of their low energy and weak coupling, detectors with low energy threshold and low background rates are mandatory. Both requirements are met by an X-ray detector based on the combination of a Micromegas gas amplification stage with a highly integrated pixel chip which allows to make full use of the Micromegas structure's granularity. It has been demonstrated that these devices can detect even single electrons. Thus, allowing for a topological background suppression as well as for detection of low energy X-ray photons creating only very few primary electrons. After the detection threshold had been evaluated to be low enough to allow for the detection of the carbon K{sub α} line at 277 eV, the detector was mounted at one of CAST's X-ray telescopes and installed along with its infrastructure in 2014. During data taking until end of 2015 background rates of less than 10{sup -4} keV/(cm{sup 2}.s) have been achieved below 2 keV. First preliminary results of the ongoing chameleon analysis and possibly an improved limit for solar chameleons are presented.

  11. Film replacement by digital x-ray detectors - the correct procedure and equipment

    International Nuclear Information System (INIS)

    Ewert, U.; Zscherpel, U.; Bavendiek, K.

    2004-01-01

    New digital detectors were developed for medical applications, which have the potential to substitute the X-ray film and revolutionise the radiological technique. Digital Detector Arrays (DDA: Flat Panel Detectors, Line Detectors) and Imaging Plates (Computed Radiography) allow a fast detection of radiographic images in a shorter time and with higher dynamic than film applications. Companies report about a reduction of exposure time down to 5 - 25% in comparison to NDT film exposures. This provides together with the reduction of consumables economical (and also ecological) benefits and short amortisation periods. But this does not always provide the same image quality as NDT film. The requirements of the European and USA standards for film radiography are analysed to derive correct requirements for the digital image quality and procedures for prediction and measurement of image quality. Basically the USA standards seem to be more tolerant for these new innovative technologies. New standard proposals use signal/noise ratio and unsharpness as dominant parameters for image quality. Specialised measurement procedures are described. The properties of the new detectors can be controlled by electronics and exposure conditions. New names appear in literature like 'direct radiography' and 'film replacement techniques'. The basic advantage of the new digital techniques is the possibility to use numeric procedures for image interpretation. Industrial radiology can be optimised for crack detection as well as for analysis of flaw depth and shape measurement. Automated flaw detection, measurement of part dimensions and detection of completeness are used for serial part inspection devices. Parallel to the development of DDA's, an extraordinary increase of Computed Tomography (CT) applications can be observed. (author)

  12. Performance evaluation of flat panel detector in x-ray fluoroscopy

    International Nuclear Information System (INIS)

    Grewal, R.K.; Mclean, I.D.

    2004-01-01

    increases the input dose during magnification to maintain image quality in much the same way as in a conventional fluoro system. Flat panel detectors in x-ray fluoroscopy systems can provide similar or better images than achievable with conventional IIs. There is no obvious distortion, typical patient dose is low and image quality is good. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine

  13. The background counting rates in a balloon borne hard X-ray telescope

    International Nuclear Information System (INIS)

    Dean, A.J.; Dipper, N.A.; Lewis, R.A.

    1986-01-01

    A detailed Monte Carlo model of a hard (20-300 keV) X-ray astronomical telescope has been developed in order to calculate the energy loss distribution of the unwanted background noise events in the prime detection elements. The spectral distributions of the background rates measured at balloon altitudes over Palestine, Texas are compared to the predictions of the theoretical model. Good agreement has been found in terms of both the overall intensity level as well as the spectral distribution. (orig.)

  14. Thin-film-based scintillators for hard x-ray microimaging detectors: the ScinTAX Project

    Science.gov (United States)

    Rack, A.; Cecilia, A.; Douissard, P.-A.; Dupré, K.; Wesemann, V.; Baumbach, T.; Couchaud, M.; Rochet, X.; Riesemeier, H.; Radtke, M.; Martin, T.

    2014-09-01

    The project ScinTAX developed novel thin scintillating films for the application in high performance X-ray imaging and subsequent introduced new X-ray detectors to the market. To achieve this aim lutetium orthosilicate (LSO) scintillators doped with different activators were grown successfully by liquid phase epitaxy. The high density of LSO (7.4 g/cm3), the effective atomic number (65.2) and the high light yield make this scintillator highly applicable for indirect X-ray detection in which the ionizing radiation is converted into visible light and then registered by a digital detector. A modular indirect detection system has been developed to fully exploit the potential of this thin film scintillator for radiographic and tomographic imaging. The system is compatible for high-resolution imaging with moderate dose as well as adaptable to intense high-dose applications where radiation hard microimaging detectors are required. This proceedings article shall review the achieved performances and technical details on this high-resolution detector system which is now available. A selected example application demonstrates the great potential of the optimized detector system for hard X-ray microimaging, i.e. either to improve image contrast due to the availability of efficient thin crystal films or to reduce the dose to the sample.

  15. Infrared LED Enhanced Spectroscopic CdZnTe Detector Working under High Fluxes of X-rays

    Directory of Open Access Journals (Sweden)

    Jakub Pekárek

    2016-09-01

    Full Text Available This paper describes an application of infrared light-induced de-polarization applied on a polarized CdZnTe detector working under high radiation fluxes. We newly demonstrate the influence of a high flux of X-rays and simultaneous 1200-nm LED illumination on the spectroscopic properties of a CdZnTe detector. CdZnTe detectors operating under high radiation fluxes usually suffer from the polarization effect, which occurs due to a screening of the internal electric field by a positive space charge caused by photogenerated holes trapped at a deep level. Polarization results in the degradation of detector charge collection efficiency. We studied the spectroscopic behavior of CdZnTe under various X-ray fluxes ranging between 5 × 10 5 and 8 × 10 6 photons per mm 2 per second. It was observed that polarization occurs at an X-ray flux higher than 3 × 10 6 mm − 2 ·s − 1 . Using simultaneous illumination of the detector by a de-polarizing LED at 1200 nm, it was possible to recover X-ray spectra originally deformed by the polarization effect.

  16. Evaluation of the x-ray response of a position-sensitive microstrip detector with an integrated readout chip

    International Nuclear Information System (INIS)

    Rossington, C.; Jaklevic, J.; Haber, C.; Spieler, H.; Reid, J.

    1990-08-01

    The performance of an SVX silicon microstrip detector and its compatible integrated readout chip have been evaluated in response to Rh Kα x-rays (average energy 20.5 keV). The energy and spatial discrimination capabilities, efficient data management and fast readout rates make it an attractive alternative to the CCD and PDA detectors now being offered for x-ray position sensitive diffraction and EXAFS work. The SVX system was designed for high energy physics applications and thus further development of the existing system is required to optimize it for use in practical x-ray experiments. For optimum energy resolution the system noise must be decreased to its previously demonstrated low levels of 2 keV FWHM at 60 keV or less, and the data handling rate of the computer must be increased. New readout chips are now available that offer the potential of better performance. 15 refs., 7 figs

  17. X-ray absorption spectroscopy of diluted system by undulator photon source and multi-element solid-state detector

    CERN Document Server

    Tanida, H

    2001-01-01

    In order to measure the extended X-ray absorption fine structure (EXAFS) spectrum of an ultra-diluted system, an optics and detector control system for a synchrotron radiation beamline is developed. The undulator gap width is continuously tuned to obtain the maximum X-ray photon flux during the energy scan for the EXAFS measurement. A piezoelectric translator optimizes the parallelism of the double crystal in a monochromator at each measurement point to compensate for mechanical errors of the monochromator, resulting in a smooth and intense X-ray photon flux during the measurement. For a detection of a weak fluorescence signal from diluted samples, a 19-element solid-state detector and digital signal processor are used. A K-edge EXAFS spectrum of iron in a myoglobin aqueous solution with a concentration of 5.58 parts per million was obtained by this system.

  18. CCD[charge-coupled device]-based synchrotron x-ray detector for protein crystallography: Performance projected from an experiment

    International Nuclear Information System (INIS)

    Strauss, M.G.; Naday, I.; Sherman, I.S.; Kraimer, M.R.; Westbrook, E.M.

    1986-01-01

    The intense x radiation from a synchrotron source could, with a suitable detector, provide a complete set of diffraction images from a protein crystal before the crystal is damaged by radiation (2 to 3 min). An area detector consisting of a 40 mm dia. x-ray fluorescing phosphor, coupled with an image intensifier and lens to a CCD image sensor, was developed to determine the effectiveness of such a detector in protein crystallography. The detector was used in an experiment with a rotating anode x-ray generator. Diffraction patterns from a lysozyme crystal obtained with this detector are compared to those obtained with film. The two images appear to be virtually identical. The flux of 10 4 x-ray photons/s was observed on the detector at the rotating anode generator. At the 6-GeV synchrotron being designed at Argonne, the flux on an 80 x 80 mm 2 detector is expected to be >10 9 photons/s. The projected design of such a synchrotron detector shows that a diffraction-peak count >10 6 could be obtained in ∼0.5 s. With an additional ∼0.5 s readout time of a 512 x 512 pixel CCD, the data acquisition time per frame would be ∼1 s so that ninety 1 0 diffraction images could be obtained, with approximately 1% precision, in less than 3 min

  19. Gallium Arsenide detectors for X-ray and electron (beta particle) spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Lioliou, G.; Barnett, A.M.

    2016-11-11

    Results characterizing GaAs p{sup +}-i-n{sup +} mesa photodiodes with a 10 µm i layer for their spectral response under illumination of X-rays and beta particles are presented. A total of 22 devices, having diameters of 200 µm and 400 µm, were electrically characterized at room temperature. All devices showed comparable characteristics with a measured leakage current ranging from 4 nA/cm{sup 2} to 67 nA/cm{sup 2} at an internal electric field of 50 kV/cm. Their unintentionally doped i layers were found to be almost fully depleted at 0 V due to their low doping density. {sup 55}Fe X-ray spectra were obtained using one 200 µm diameter device and one 400 µm diameter device. The best energy resolution (FWHM at 5.9 keV) achieved was 625 eV using the 200 µm and 740 eV using the 400 µm diameter device, respectively. Noise analysis showed that the limiting factor for the energy resolution of the system was the dielectric noise; if this noise was eliminated by better design of the front end of the readout electronics, the achievable resolution would be 250 eV. {sup 63}Ni beta particle spectra obtained using the 200 µm diameter device showed the potential utility of these detectors for electron and beta particle detection. The development of semiconductor electron spectrometers is important particularly for space plasma physics; such devices may find use in future space missions to study the plasma environment of Jupiter and Europa and the predicted electron impact excitation of water vapor plumes from Europa hypothesized as a result of recent Hubble Space Telescope (HST) UV observations.

  20. Gallium Arsenide detectors for X-ray and electron (beta particle) spectroscopy

    Science.gov (United States)

    Lioliou, G.; Barnett, A. M.

    2016-11-01

    Results characterizing GaAs p+-i-n+ mesa photodiodes with a 10 μm i layer for their spectral response under illumination of X-rays and beta particles are presented. A total of 22 devices, having diameters of 200 μm and 400 μm, were electrically characterized at room temperature. All devices showed comparable characteristics with a measured leakage current ranging from 4 nA/cm2 to 67 nA/cm2 at an internal electric field of 50 kV/cm. Their unintentionally doped i layers were found to be almost fully depleted at 0 V due to their low doping density. 55Fe X-ray spectra were obtained using one 200 μm diameter device and one 400 μm diameter device. The best energy resolution (FWHM at 5.9 keV) achieved was 625 eV using the 200 μm and 740 eV using the 400 μm diameter device, respectively. Noise analysis showed that the limiting factor for the energy resolution of the system was the dielectric noise; if this noise was eliminated by better design of the front end of the readout electronics, the achievable resolution would be 250 eV. 63Ni beta particle spectra obtained using the 200 μm diameter device showed the potential utility of these detectors for electron and beta particle detection. The development of semiconductor electron spectrometers is important particularly for space plasma physics; such devices may find use in future space missions to study the plasma environment of Jupiter and Europa and the predicted electron impact excitation of water vapor plumes from Europa hypothesized as a result of recent Hubble Space Telescope (HST) UV observations.

  1. EIGER: Next generation single photon counting detector for X-ray applications

    Energy Technology Data Exchange (ETDEWEB)

    Dinapoli, Roberto, E-mail: roberto.dinapoli@psi.ch [Paul Scherrer Institut, 5232 Villigen PSI (Switzerland); Bergamaschi, Anna; Henrich, Beat; Horisberger, Roland; Johnson, Ian; Mozzanica, Aldo; Schmid, Elmar; Schmitt, Bernd; Schreiber, Akos; Shi, Xintian; Theidel, Gerd [Paul Scherrer Institut, 5232 Villigen PSI (Switzerland)

    2011-09-11

    EIGER is an advanced family of single photon counting hybrid pixel detectors, primarily aimed at diffraction experiments at synchrotrons. Optimization of maximal functionality and minimal pixel size (using a 0.25{mu}m process and conserving the radiation tolerant design) has resulted in 75x75{mu}m{sup 2} pixels. Every pixel comprises a preamplifier, shaper, discriminator (with a 6 bit DAC for threshold trimming), a configurable 4/8/12 bit counter with double buffering, as well as readout, control and test circuitry. A novel feature of this chip is its double buffered counter, meaning a next frame can be acquired while the previous one is being readout. An array of 256x256 pixels fits on a {approx}2x2cm{sup 2} chip and a sensor of {approx}8x4cm{sup 2} will be equipped with eight readout chips to form a module containing 0.5 Mpixel. Several modules can then be tiled to form larger area detectors. Detectors up to 4x8 modules (16 Mpixel) are planned. To achieve frame rates of up to 24 kHz the readout architecture is highly parallel, and the chip readout happens in parallel on 32 readout lines with a 100 MHz Double Data Rate clock. Several chips and singles (i.e. a single chip bump-bonded to a single chip silicon sensor) were tested both with a lab X-ray source and at Swiss Light Source (SLS) beamlines. These tests demonstrate the full functionality of the chip and provide a first assessment of its performance. High resolution X-ray images and 'high speed movies' were produced, even without threshold trimming, at the target system frame rates (up to {approx}24kHz in 4 bit mode). In parallel, dedicated hardware, firmware and software had to be developed to comply with the enormous data rate the chip is capable of delivering. Details of the chip design and tests will be given, as well as highlights of both test and final readout systems.

  2. X-ray instrumentation in astronomy

    International Nuclear Information System (INIS)

    Cuhlane, J.L.

    1985-01-01

    This book presents the proceedings of a conference devoted to x-ray instrumentation in astronomy. Special sections are: AXAF X-Ray Optical Systems; Specialized X-Ray Systems; X-Ray Optical Systems I; X-Ray Optical Systems II; Gas Filled X-Ray Detectors II; The NASA Advanced X-Ray Astrophysics Facility; X-Ray and EUV Spectrometers; Microchannel Plates; and Solid State Detectors

  3. Research on multi-spectrum detector in high-energy dual-energy X-ray imaging system

    International Nuclear Information System (INIS)

    Li Qinghua; Wang Xuewu; Li Jianmin; Kang Kejun; Li Yuanjing; Zhong Huaqiang

    2008-01-01

    The high-energy dual-energy X-ray imaging system can discriminate the material of the objects inspected, but when the objects are too thin, the discrimination becomes very difficult. This paper proposes the use of multi-spectrum detector to improve the ability to discriminate thin material, and a series of simulation were done with the Monte Carlo method. Firstly the X-ray depositions in the detectors with different thickness were calculated, and then the discrimination effects with different detector structure and parameters were calculated. The simulation results validated that using appropriate multi-spectrum detector can improve the discrimination accuracy of thin material, particularly thin high-Z material. (authors)

  4. Simultaneous acquisition of X-ray spectra using a multi-wire, position-sensitive gas flow detector

    International Nuclear Information System (INIS)

    Beaven, Peter A.; Marmotti, Mauro; Kampmann, Reinhard; Knoth, Joachim; Schwenke, Heinrich

    2003-01-01

    A multi-wire, gas-filled position-sensitive detector has been developed for the simultaneous recording of wavelength-dispersed X-ray signals that enables X-ray fluorescence spectrometry with a limited multi-element capability in the low Z element range. Details of the modular construction of the detector are given. The detector performance was characterized using Al-Kα radiation and a variable slit system. The detector has been applied in a laboratory spectrometer equipped with an electron source and a double multilayer mirror device as the wavelength-dispersing element. Spectra from Al and Si obtained in the simultaneous acquisition mode show good agreement with calculations performed using a ray-tracing model

  5. Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype.

    Science.gov (United States)

    Speidel, Michael A; Tomkowiak, Michael T; Raval, Amish N; Dunkerley, David A P; Slagowski, Jordan M; Kahn, Paul; Ku, Jamie; Funk, Tobias

    Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system for low dose cardiac imaging. The use of a narrow scanned x-ray beam in SBDX reduces detected x-ray scatter and improves dose efficiency, however the tight beam collimation also limits the maximum achievable x-ray fluence. To increase the fluence available for imaging, we have constructed a new SBDX prototype with a wider x-ray beam, larger-area detector, and new real-time image reconstructor. Imaging is performed with a scanning source that generates 40,328 narrow overlapping projections from 71 × 71 focal spot positions for every 1/15 s scan period. A high speed 2-mm thick CdTe photon counting detector was constructed with 320×160 elements and 10.6 cm × 5.3 cm area (full readout every 1.28 μs), providing an 86% increase in area over the previous SBDX prototype. A matching multihole collimator was fabricated from layers of tungsten, brass, and lead, and a multi-GPU reconstructor was assembled to reconstruct the stream of captured detector images into full field-of-view images in real time. Thirty-two tomosynthetic planes spaced by 5 mm plus a multiplane composite image are produced for each scan frame. Noise equivalent quanta on the new SBDX prototype measured 63%-71% higher than the previous prototype. X-ray scatter fraction was 3.9-7.8% when imaging 23.3-32.6 cm acrylic phantoms, versus 2.3-4.2% with the previous prototype. Coronary angiographic imaging at 15 frame/s was successfully performed on the new SBDX prototype, with live display of either a multiplane composite or single plane image.

  6. LAMBDA 2M GaAs—A multi-megapixel hard X-ray detector for synchrotrons

    Science.gov (United States)

    Pennicard, D.; Smoljanin, S.; Pithan, F.; Sarajlic, M.; Rothkirch, A.; Yu, Y.; Liermann, H. P.; Morgenroth, W.; Winkler, B.; Jenei, Z.; Stawitz, H.; Becker, J.; Graafsma, H.

    2018-01-01

    Synchrotrons can provide very intense and focused X-ray beams, which can be used to study the structure of matter down to the atomic scale. In many experiments, the quality of the results depends strongly on detector performance; in particular, experiments studying dynamics of samples require fast, sensitive X-ray detectors. "LAMBDA" is a photon-counting hybrid pixel detector system for experiments at synchrotrons, based on the Medipix3 readout chip. Its main features are a combination of comparatively small pixel size (55 μm), high readout speed at up to 2000 frames per second with no time gap between images, a large tileable module design, and compatibility with high-Z sensors for efficient detection of higher X-ray energies. A large LAMBDA system for hard X-ray detection has been built using Cr-compensated GaAs as a sensor material. The system is composed of 6 GaAs tiles, each of 768 by 512 pixels, giving a system with approximately 2 megapixels and an area of 8.5 by 8.5 cm2. While the sensor uniformity of GaAs is not as high as that of silicon, its behaviour is stable over time, and it is possible to correct nonuniformities effectively by postprocessing of images. By using multiple 10 Gigabit Ethernet data links, the system can be read out at the full speed of 2000 frames per second. The system has been used in hard X-ray diffraction experiments studying the structure of samples under extreme pressure in diamond anvil cells. These experiments can provide insight into geological processes. Thanks to the combination of high speed readout, large area and high sensitivity to hard X-rays, it is possible to obtain previously unattainable information in these experiments about atomic-scale structure on a millisecond timescale during rapid changes of pressure or temperature.

  7. CdZnTe Image Detectors for Hard-X-Ray Telescopes

    Science.gov (United States)

    Chen, C. M. Hubert; Cook, Walter R.; Harrison, Fiona A.; Lin, Jiao Y. Y.; Mao, Peter H.; Schindler, Stephen M.

    2005-01-01

    Arrays of CdZnTe photodetectors and associated electronic circuitry have been built and tested in a continuing effort to develop focal-plane image sensor systems for hard-x-ray telescopes. Each array contains 24 by 44 pixels at a pitch of 498 m. The detector designs are optimized to obtain low power demand with high spectral resolution in the photon- energy range of 5 to 100 keV. More precisely, each detector array is a hybrid of a CdZnTe photodetector array and an application-specific integrated circuit (ASIC) containing an array of amplifiers in the same pixel pattern as that of the detectors. The array is fabricated on a single crystal of CdZnTe having dimensions of 23.6 by 12.9 by 2 mm. The detector-array cathode is a monolithic platinum contact. On the anode plane, the contact metal is patterned into the aforementioned pixel array, surrounded by a guard ring that is 1 mm wide on three sides and is 0.1 mm wide on the fourth side so that two such detector arrays can be placed side-by-side to form a roughly square sensor area with minimal dead area between them. Figure 1 shows two anode patterns. One pattern features larger pixel anode contacts, with a 30-m gap between them. The other pattern features smaller pixel anode contacts plus a contact for a shaping electrode in the form of a grid that separates all the pixels. In operation, the grid is held at a potential intermediate between the cathode and anode potentials to steer electric charges toward the anode in order to reduce the loss of charges in the inter-anode gaps. The CdZnTe photodetector array is mechanically and electrically connected to the ASIC (see Figure 2), either by use of indium bump bonds or by use of conductive epoxy bumps on the CdZnTe array joined to gold bumps on the ASIC. Hence, the output of each pixel detector is fed to its own amplifier chain.

  8. High density processing electronics for superconducting tunnel junction x-ray detector arrays

    Energy Technology Data Exchange (ETDEWEB)

    Warburton, W.K., E-mail: bill@xia.com [XIA LLC, 31057 Genstar Road, Hayward, CA 94544 (United States); Harris, J.T. [XIA LLC, 31057 Genstar Road, Hayward, CA 94544 (United States); Friedrich, S. [Lawrence Livermore National Laboratory, Livermore, CA 94550 (United States)

    2015-06-01

    Superconducting tunnel junctions (STJs) are excellent soft x-ray (100–2000 eV) detectors, particularly for synchrotron applications, because of their ability to obtain energy resolutions below 10 eV at count rates approaching 10 kcps. In order to achieve useful solid detection angles with these very small detectors, they are typically deployed in large arrays – currently with 100+ elements, but with 1000 elements being contemplated. In this paper we review a 5-year effort to develop compact, computer controlled low-noise processing electronics for STJ detector arrays, focusing on the major issues encountered and our solutions to them. Of particular interest are our preamplifier design, which can set the STJ operating points under computer control and achieve 2.7 eV energy resolution; our low noise power supply, which produces only 2 nV/√Hz noise at the preamplifier's critical cascode node; our digital processing card that digitizes and digitally processes 32 channels; and an STJ I–V curve scanning algorithm that computes noise as a function of offset voltage, allowing an optimum operating point to be easily selected. With 32 preamplifiers laid out on a custom 3U EuroCard, and the 32 channel digital card in a 3U PXI card format, electronics for a 128 channel array occupy only two small chassis, each the size of a National Instruments 5-slot PXI crate, and allow full array control with simple extensions of existing beam line data collection packages.

  9. Filter-fluorescer x-ray spectrometer using solid state detectors for γ-ray background reduction

    International Nuclear Information System (INIS)

    Yokoi, Takashi; Kitagawa, Yoneyoshi; Shiraga, Hiroyuki; Matsunaga, Hirohide; Kato, Yoshiaki; Yamanaka, Chiyoe.

    1986-01-01

    Filter-fluorescer x-ray spectrometer using solid state photo-detectors instead of the photomultiplier tubes in order to reduce the γ-ray background noise is reported. A significant reduction of the γ-ray background noise is expected, because solid state photo-detectors are very small in size compared with the photomultiplier tubes. It has been confirmed that the γ-ray background is reduced in the target irradiation experiments with the Gekko MII glass laser. (author)

  10. SEDRX: A computer program for the simulation Si(Li) and Ge(Hp) x-ray detectors efficiency

    International Nuclear Information System (INIS)

    Benamar, M.A.; Benouali, A.; Tchantchane, A.; Azbouche, A.; Tobbeche, S. Centre de Developpement des Techniques Nucleaires, Algiers; Labo. des Techniques Nucleaires)

    1992-12-01

    The difficulties encountered in measuring the x-ray detectors efficiency has motivated to develop a computer program to simulate this parameter. this program computes the efficiency of detectors as a function of energy. the computation of this parameter is based on the fitting coefficients of absorption in the case of photoelectric, coherent and incoherent factors. These coefficients are given by Mc Master library or may be determined by the interpolation based on cubic splines

  11. Dynamic defectoscopy with flat panel and CdTe Timepix X-ray detectors combined with an optical camera

    Czech Academy of Sciences Publication Activity Database

    Vavřík, Daniel; Fauler, A.; Fiederle, M.; Jandejsek, Ivan; Jakůbek, J.; Tureček, D.; Zwerger, A.

    2013-01-01

    Roč. 8, April (2013), C04009 ISSN 1748-0221. [International Workshop on Radiation Imaging Detectors /14./. Figueira da Foz, Coimbra, 01.07.2012-05.07.2012] R&D Projects: GA ČR(CZ) GA103/09/2101 Institutional support: RVO:68378297 Keywords : X-ray digital radiography * fracture mechanics * crack path * X-ray defectoscopy Subject RIV: JM - Building Engineering Impact factor: 1.526, year: 2013 http://iopscience.iop.org/1748-0221/8/04/C04009/

  12. X-ray radiation detectors of 'scintillator-photoreceiving device type' for industrial digital radiography with improved spatial resolution

    International Nuclear Information System (INIS)

    Ryzhykov, V.D.; Lysetska, O.K.; Opolonin, O.D.; Kozin, D.N.

    2003-01-01

    Main types of photo receivers used in X-ray digital radiography systems are luminescent screens that transfer the optical image onto charge collection instruments, which require cooling, and semiconductor silicon detectors, which limit the contrast sensitivity. We have developed and produced X-ray radiation detectors of 'scintillator-photoreceiving device' (S-PRD) type, which are integrally located on the inverse side of the photodiode (PD). The receiving-converting circuit (RCC) is designed for data conversion into digital form and their input into PC. Software is provided for RCC control and image visualization. Main advantages of these detectors are high industrial resolution (3-5 line pairs per mm), detecting activity up to 20 μm, controlled sensitivity, low weight and small size, imaging low (0.1-0.3 mrad) object dose in real time. In this work, main characteristics of 32-, 64- and 1024-channel detectors of S-PRD type were studied and compared for X-ray sensitivity with S-PD detectors. Images of the tested objects have been obtained. Recommendations are given on the use of different scintillation materials, depending upon the purpose of a digital radiographic system. The detectors operate in a broad energy range of ionizing radiation, hence the size of the controlled object is not limited. The system is sufficiently powerful to ensure frontal (through two walls) observation of pipelines with wall thickness up to 10 cm

  13. Monte Carlo simulation of the response functions of Cd Te detectors to be applied in X-rays spectroscopy

    International Nuclear Information System (INIS)

    Tomal, A.; Lopez G, A. H.; Santos, J. C.; Costa, P. R.

    2014-08-01

    In this work, the energy response functions of a Cd Te detector were obtained by Monte Carlo simulation in the energy range from 5 to 150 keV, using the Penelope code. The response functions simulated included the finite detector resolution and the carrier transport. The simulated energy response matrix was validated through comparison with experimental results obtained for radioactive sources. In order to investigate the influence of the correction by the detector response at diagnostic energy range, x-ray spectra were measured using a Cd Te detector (model Xr-100-T, Amptek), and then corrected by the energy response of the detector using the stripping procedure. Results showed that the Cd Te exhibit good energy response at low energies (below 40 keV), showing only small distortions on the measured spectra. For energies below about 70 keV, the contribution of the escape of Cd- and Te-K x-rays produce significant distortions on the measured x-ray spectra. For higher energies, the most important correction is the detector efficiency and the carrier trapping effects. The results showed that, after correction by the energy response, the measured spectra are in good agreement with those provided by different models from the literature. Finally, our results showed that the detailed knowledge of the response function and a proper correction procedure are fundamental for achieve more accurate spectra from which several qualities parameters (i.e. half-value layer, effective energy and mean energy) can be determined. (Author)

  14. Monte Carlo simulation of the response functions of Cd Te detectors to be applied in X-rays spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Tomal, A. [Universidade Federale de Goias, Instituto de Fisica, Campus Samambaia, 74001-970, Goiania, (Brazil); Lopez G, A. H.; Santos, J. C.; Costa, P. R., E-mail: alessandra_tomal@yahoo.com.br [Universidade de Sao Paulo, Instituto de Fisica, Rua du Matao Travessa R. 187, Cidade Universitaria, 05508-090 Sao Paulo (Brazil)

    2014-08-15

    In this work, the energy response functions of a Cd Te detector were obtained by Monte Carlo simulation in the energy range from 5 to 150 keV, using the Penelope code. The response functions simulated included the finite detector resolution and the carrier transport. The simulated energy response matrix was validated through comparison with experimental results obtained for radioactive sources. In order to investigate the influence of the correction by the detector response at diagnostic energy range, x-ray spectra were measured using a Cd Te detector (model Xr-100-T, Amptek), and then corrected by the energy response of the detector using the stripping procedure. Results showed that the Cd Te exhibit good energy response at low energies (below 40 keV), showing only small distortions on the measured spectra. For energies below about 70 keV, the contribution of the escape of Cd- and Te-K x-rays produce significant distortions on the measured x-ray spectra. For higher energies, the most important correction is the detector efficiency and the carrier trapping effects. The results showed that, after correction by the energy response, the measured spectra are in good agreement with those provided by different models from the literature. Finally, our results showed that the detailed knowledge of the response function and a proper correction procedure are fundamental for achieve more accurate spectra from which several qualities parameters (i.e. half-value layer, effective energy and mean energy) can be determined. (Author)

  15. Monte Carlo semi-empirical model for Si(Li) x-ray detector: Differences between nominal and fitted parameters

    Energy Technology Data Exchange (ETDEWEB)

    Lopez-Pino, N.; Padilla-Cabal, F.; Garcia-Alvarez, J. A.; Vazquez, L.; D' Alessandro, K.; Correa-Alfonso, C. M. [Departamento de Fisica Nuclear, Instituto Superior de Tecnologia y Ciencias Aplicadas (InSTEC) Ave. Salvador Allende y Luaces. Quinta de los Molinos. Habana 10600. A.P. 6163, La Habana (Cuba); Godoy, W.; Maidana, N. L.; Vanin, V. R. [Laboratorio do Acelerador Linear, Instituto de Fisica - Universidade de Sao Paulo Rua do Matao, Travessa R, 187, 05508-900, SP (Brazil)

    2013-05-06

    A detailed characterization of a X-ray Si(Li) detector was performed to obtain the energy dependence of efficiency in the photon energy range of 6.4 - 59.5 keV, which was measured and reproduced by Monte Carlo (MC) simulations. Significant discrepancies between MC and experimental values were found when the manufacturer parameters of the detector were used in the simulation. A complete Computerized Tomography (CT) detector scan allowed to find the correct crystal dimensions and position inside the capsule. The computed efficiencies with the resulting detector model differed with the measured values no more than 10% in most of the energy range.

  16. Development of Small-Pixel CZT Detectors for Future High-Resolution Hard X-ray Missions

    Science.gov (United States)

    Beilicke, Matthias

    Owing to recent breakthroughs in grazing incidence mirror technology, next-generation hard X-ray telescopes will achieve angular resolutions of between 5 and 10 arc seconds - about an order of magnitude better than that of the NuSTAR hard X-ray telescope. As a consequence, the next generation of hard X-ray telescopes will require pixelated hard X- ray detectors with pixels on a grid with a lattice constant of between 120 and 240 um. Additional detector requirements include a low energy threshold of less than 5 keV and an energy resolution of less than 1 keV. The science drivers for a high angular-resolution hard X-ray mission include studies and measurements of black hole spins, the cosmic evolution of super-massive black holes, AGN feedback, and the behavior of matter at very high densities. We propose a R&D research program to develop, optimize and study the performance of 100-200 um pixel pitch CdTe and Cadmium Zinc Telluride (CZT) detectors of 1-2 mm thickness. Our program aims at a comparison of the performance achieved with CdTe and CZT detectors, and the optimization of the pixel, steering grid, and guard ring anode patterns. Although these studies will use existing ASICs (Application Specific Integrated Circuits), our program also includes modest funds for the development of an ultra-low noise ASIC with a 2-D grid of readout pads that can be directly bonded to the 100-200 um pixel pitch CdTe and CZT detectors. The team includes the Washington University group (Prof. M. Beilicke and Co-I Prof. H.S.W. Krawczynski et al.), and co-investigator G. De Geronimo at Brookhaven National Laboratory (BNL). The Washington University group has a 10 year track record of innovative CZT detector R&D sponsored by the NASA Astronomy and Physics Research and Analysis (APRA) program. The accomplishments to date include the development of CZT detectors with pixel pitches between 350 um and 2.5 mm for the ProtoExist, EXIST, and X-Calibur hard X-ray missions with some of the best

  17. Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging.

    Science.gov (United States)

    Tanaka, Rie

    2016-07-01

    Dynamic chest radiography is a flat-panel detector (FPD)-based functional X-ray imaging, which is performed as an additional examination in chest radiography. The large field of view (FOV) of FPDs permits real-time observation of the entire lungs and simultaneous right-and-left evaluation of diaphragm kinetics. Most importantly, dynamic chest radiography provides pulmonary ventilation and circulation findings as slight changes in pixel value even without the use of contrast media; the interpretation is challenging and crucial for a better understanding of pulmonary function. The basic concept was proposed in the 1980s; however, it was not realized until the 2010s because of technical limitations. Dynamic FPDs and advanced digital image processing played a key role for clinical application of dynamic chest radiography. Pulmonary ventilation and circulation can be quantified and visualized for the diagnosis of pulmonary diseases. Dynamic chest radiography can be deployed as a simple and rapid means of functional imaging in both routine and emergency medicine. Here, we focus on the evaluation of pulmonary ventilation and circulation. This review article describes the basic mechanism of imaging findings according to pulmonary/circulation physiology, followed by imaging procedures, analysis method, and diagnostic performance of dynamic chest radiography.

  18. Edgeless silicon sensors for Medipix-based large-area X-ray imaging detectors

    International Nuclear Information System (INIS)

    Bosma, M J; Visser, J; Koffeman, E N; Evrard, O; De Moor, P; De Munck, K; Tezcan, D Sabuncuoglu

    2011-01-01

    Some X-ray imaging applications demand sensitive areas exceeding the active area of a single sensor. This requires a seamless tessellation of multiple detector modules with edgeless sensors. Our research is aimed at minimising the insensitive periphery that isolates the active area from the edge. Reduction of the edge-defect induced charge injection, caused by the deleterious effects of dicing, is an important step. We report on the electrical characterisation of 300 μm thick edgeless silicon p + -ν-n + diodes, diced using deep reactive ion etching. Sensors with both n-type and p-type stop rings were fabricated in various edge topologies. Leakage currents in the active area are compared with those of sensors with a conventional design. As expected, we observe an inverse correlation between leakage-current density and both the edge distance and stop-ring width. From this correlation we determine a minimum acceptable edge distance of 50 μm. We also conclude that structures with a p-type stop ring show lower leakage currents and higher breakdown voltages than the ones with an n-type stop ring.

  19. High-accuracy X-ray detector calibration based on cryogenic radiometry

    Science.gov (United States)

    Krumrey, M.; Cibik, L.; Müller, P.

    2010-06-01

    Cryogenic electrical substitution radiometers (ESRs) are absolute thermal detectors, based on the equivalence of electrical power and radiant power. Their core piece is a cavity absorber, which is typically made of copper to achieve a short response time. At higher photon energies, the use of copper prevents the operation of ESRs due to increasing transmittance. A new absorber design for hard X-rays has been developed at the laboratory of the Physikalisch-Technische Bundesanstalt (PTB) at the electron storage ring BESSY II. The Monte Carlo simulation code Geant4 was applied to optimize its absorptance for photon energies of up to 60 keV. The measurement of the radiant power of monochromatized synchrotron radiation was achieved with relative standard uncertainties of less than 0.2 %, covering the entire photon energy range of three beamlines from 50 eV to 60 keV. Monochromatized synchrotron radiation of high spectral purity is used to calibrate silicon photodiodes against the ESR for photon energies up to 60 keV with relative standard uncertainties below 0.3 %. For some silicon photodiodes, the photocurrent is not linear with the incident radiant power.

  20. High-accuracy X-ray detector calibration based on cryogenic radiometry

    International Nuclear Information System (INIS)

    Krumrey, M.; Cibik, L.; Mueller, P.

    2010-01-01

    Cryogenic electrical substitution radiometers (ESRs) are absolute thermal detectors, based on the equivalence of electrical power and radiant power. Their core piece is a cavity absorber, which is typically made of copper to achieve a short response time. At higher photon energies, the use of copper prevents the operation of ESRs due to increasing transmittance. A new absorber design for hard X-rays has been developed at the laboratory of the Physikalisch-Technische Bundesanstalt (PTB) at the electron storage ring BESSY II. The Monte Carlo simulation code Geant4 was applied to optimize its absorptance for photon energies of up to 60 keV. The measurement of the radiant power of monochromatized synchrotron radiation was achieved with relative standard uncertainties of less than 0.2 %, covering the entire photon energy range of three beamlines from 50 eV to 60 keV. Monochromatized synchrotron radiation of high spectral purity is used to calibrate silicon photodiodes against the ESR for photon energies up to 60 keV with relative standard uncertainties below 0.3 %. For some silicon photodiodes, the photocurrent is not linear with the incident radiant power.

  1. Development of a 32-channel ASIC for an X-ray APD detector onboard the ISS

    Science.gov (United States)

    Arimoto, Makoto; Harita, Shohei; Sugita, Satoshi; Yatsu, Yoichi; Kawai, Nobuyuki; Ikeda, Hirokazu; Tomida, Hiroshi; Isobe, Naoki; Ueno, Shiro; Mihara, Tatehiro; Serino, Motoko; Kohmura, Takayoshi; Sakamoto, Takanori; Yoshida, Atsumasa; Tsunemi, Hiroshi; Hatori, Satoshi; Kume, Kyo; Hasegawa, Takashi

    2018-02-01

    We report on the design and performance of a mixed-signal application specific integrated circuit (ASIC) dedicated to avalanche photodiodes (APDs) in order to detect hard X-ray emissions in a wide energy band onboard the International Space Station. To realize wide-band detection from 20 keV to 1 MeV, we use Ce:GAGG scintillators, each coupled to an APD, with low-noise front-end electronics capable of achieving a minimum energy detection threshold of 20 keV. The developed ASIC has the ability to read out 32-channel APD signals using 0.35 μm CMOS technology, and an analog amplifier at the input stage is designed to suppress the capacitive noise primarily arising from the large detector capacitance of the APDs. The ASIC achieves a performance of 2099 e- + 1.5 e-/pF at root mean square (RMS) with a wide 300 fC dynamic range. Coupling a reverse-type APD with a Ce:GAGG scintillator, we obtain an energy resolution of 6.7% (FWHM) at 662 keV and a minimum detectable energy of 20 keV at room temperature (20 °C). Furthermore, we examine the radiation tolerance for space applications by using a 90 MeV proton beam, confirming that the ASIC is free of single-event effects and can operate properly without serious degradation in analog and digital processing.

  2. POLAR: A Space-borne X-Ray Polarimeter for Transient Sources

    Science.gov (United States)

    Orsi, S.; Polar Collaboration

    2011-02-01

    POLAR is a novel compact Compton X-ray polarimeter designed to measure the linear polarization of the prompt emission of Gamma Ray Bursts (GRB) and other strong transient sources such as soft gamma repeaters and solar flares in the energy range 50-500 keV. A detailed measurement of the polarization from astrophysical sources will lead to a better understanding of the source geometry and emission mechanisms. POLAR is expected to observe every year several GRBs with a minimum detectable polarization smaller than 10%, thanks to its large modulation factor, effective area, and field of view. POLAR consists of 1600 low-Z plastic scintillator bars, divided in 25 independent modular units, each read out by one flat-panel multi-anode photomultiplier. The design of POLAR is reviewed, and results of tests of one modular unit of the engineering and qualification model (EQM) of POLAR with synchrotron radiation are presented. After construction and testing of the full EQM, we will start building the flight model in 2011, in view of the launch foreseen in 2013.

  3. In situ micro-focused X-ray beam characterization with a lensless camera using a hybrid pixel detector

    International Nuclear Information System (INIS)

    Kachatkou, Anton; Marchal, Julien; Silfhout, Roelof van

    2014-01-01

    Position and size measurements of a micro-focused X-ray beam, using an X-ray beam imaging device based on a lensless camera that collects radiation scattered from a thin foil placed in the path of the beam at an oblique angle, are reported. Results of studies on micro-focused X-ray beam diagnostics using an X-ray beam imaging (XBI) instrument based on the idea of recording radiation scattered from a thin foil of a low-Z material with a lensless camera are reported. The XBI instrument captures magnified images of the scattering region within the foil as illuminated by the incident beam. These images contain information about beam size, beam position and beam intensity that is extracted during dedicated signal processing steps. In this work the use of the device with beams for which the beam size is significantly smaller than that of a single detector pixel is explored. The performance of the XBI device equipped with a state-of-the-art hybrid pixel X-ray imaging sensor is analysed. Compared with traditional methods such as slit edge or wire scanners, the XBI micro-focused beam characterization is significantly faster and does not interfere with on-going experiments. The challenges associated with measuring micrometre-sized beams are described and ways of optimizing the resolution of beam position and size measurements of the XBI instrument are discussed

  4. arXiv Energy Dependent Features of X-ray Signals in a GridPix Detector

    CERN Document Server

    Krieger, Christoph; Vafeiadis, Theodoros; Desch, Klaus

    2018-06-11

    We report on the calibration of an argon/isobutane (97.7%/2.3%)-filled GridPix detector with soft X-rays (277 eV to 8 keV) using the variable energy X-ray source of the CAST Detector Lab at CERN. We study the linearity and energy resolution of the detector using both the number of pixels hit and the total measured charge as energy measures. For the latter, the energy resolution σE∕E is better than 10% (20%) for energies above 2 keV (0.5 keV). Several characteristics of the recorded events are studied.

  5. Gaining efficiency and resolution in soft X-ray emission spectrometers thanks to directly illuminated CCD detectors

    International Nuclear Information System (INIS)

    Dinardo, M.E.; Piazzalunga, A.; Braicovich, L.; Bisogni, V.; Dallera, C.; Giarda, K.; Marcon, M.; Tagliaferri, A.; Ghiringhelli, G.

    2007-01-01

    The back-illuminated charge coupled devices (CCD) are suitable for soft X-ray photon detection. Their nominal performances suggest that they can boost both efficiency and resolving power of X-ray spectrometers based on diffraction gratings and two-dimensional position sensitive detectors. We tested the performances of two commercially available CCDs, intended to replace a more traditional microchannel plate (MCP) detector. Our tests show that the devices have excellent performances in terms of dark current, response linearity, detection efficiency and spatial resolution. We observed that the CCDs have better efficiency (more than 10 times) and better resolution (∼3 times) than the MCP. Moreover we found an intrinsic limit for the spatial resolution, which is almost independent of the detector pixel size and is estimated around 25 μm

  6. Compression of Born ratio for fluorescence molecular tomography/x-ray computed tomography hybrid imaging: methodology and in vivo validation.

    Science.gov (United States)

    Mohajerani, Pouyan; Ntziachristos, Vasilis

    2013-07-01

    The 360° rotation geometry of the hybrid fluorescence molecular tomography/x-ray computed tomography modality allows for acquisition of very large datasets, which pose numerical limitations on the reconstruction. We propose a compression method that takes advantage of the correlation of the Born-normalized signal among sources in spatially formed clusters to reduce the size of system model. The proposed method has been validated using an ex vivo study and an in vivo study of a nude mouse with a subcutaneous 4T1 tumor, with and without inclusion of a priori anatomical information. Compression rates of up to two orders of magnitude with minimum distortion of reconstruction have been demonstrated, resulting in large reduction in weight matrix size and reconstruction time.

  7. A 3D CZT hard x-ray polarimeter for a balloon-borne payload

    DEFF Research Database (Denmark)

    Caroli, E.; Alvarez, J. M.; Auricchio, N.

    2012-01-01

    be optimized also for this type of measurement. In this framework, we present the concept of a small high-performance spectrometer designed for polarimetry between 100 and 1000 keV suitable as a stratospheric balloon-borne payload dedicated to perform an accurate and reliable measurement of the polarization...

  8. Calorimetric low - temperature detectors for high resolution X-ray spectroscopy on stored highly stripped heavy ions

    International Nuclear Information System (INIS)

    Bleile, A.; Egelhof, P.; Kraft, S.; Meier, H.J.; Shrivastava, A.; Weber, M.; McCammon, D.; Stahle, C.K.

    2001-09-01

    The accurate determination of the Lamb shift in heavy hydrogen-like ions provides a sensitive test of quantum electrodynamics in very strong Coulomb fields, not accessible otherwise. For the investigation of the Lyman-α transitions in 208 Pb 81+ or 238 U 91+ with sufficient accuracy, a high resolution calorimetric detector for hard X-rays (E ≤ 100 keV) is presently being developed. The detector modules consist of arrays of silicon thermistors and of X-ray absorbers made of high-Z material to optimize the absorption efficiency. The detectors are housed in a specially designed 3 He/ 4 He dilution refrigerator with a side arm which fits to the internal target geometry of the storage ring ESR at GSI Darmstadt. The detector performance presently achieved is already close to fulfill the demands of the Lamb shift experiment. For a prototype detector pixel with a 0.2 mm 2 x 47 μm Pb absorber an energy resolution of ΔE FWHM = 65 eV is obtained for 60 keV X-rays. (orig.)

  9. New medical imaging systems exploiting the energy dispersive X-ray diffraction with spectrometric CdZnTe based detector

    International Nuclear Information System (INIS)

    Barbes, Damien

    2016-01-01

    This thesis studies the interest of measuring the coherent scattering of X-rays for breast diagnosis imaging. Nowadays, most of X-ray-based medical imaging techniques use the information of X-rays attenuation through the tissues. It is the case for mammography, the most common breast imaging modality. The recent emergence of energy resolved detectors (based on semiconductors in particular) allows to consider using another phenomenon: the coherent X-ray scattering. Measurement of diffracted spectra can provide new information related to the molecular structure of the examined tissues, in order to improve their characterization and therefore improve the final diagnosis. Two modalities are considered: the breast cancer detection in vivo, following a suspicious mammography result, or biopsy analysis. The coherent scattering measurement system developed during this thesis work uses energy-resolved CdZnTe-based detectors, these detectors combining performances (energy resolution, sensitivity, spatial resolution, and compactness) promising for clinical application. This system is also based on the detector pixelation, which allows to provide an imaging modality capable of characterizing analyzed materials or tissues in one direction without any translation or rotation. A complete study of the measurement system is proposed in this thesis, structured in three main parts: modeling and simulation of the system, development of the processing of the data measured by the detector in order to image and characterize the analyzed sample and finally, designing of a new and more complex experimental setup based on a whole detector and multi-slit collimation system. An experimental validation is proposed for each of these three parts. (author) [fr

  10. GEM gas detectors for soft X-ray imaging in fusion devices with neutron–gamma background

    Energy Technology Data Exchange (ETDEWEB)

    Pacella, Danilo, E-mail: danilo.pacella@enea.it [Associazione EURATOM-ENEA, C.R. Frascati, Via E. Fermi 45, 00044 Frascati, Roma (Italy); Romano, Afra; Gabellieri, Lori [Associazione EURATOM-ENEA, C.R. Frascati, Via E. Fermi 45, 00044 Frascati, Roma (Italy); Murtas, Fabrizio [Istituto Nazionale di Fisica Nucleare, Via E. Fermi 45, 00044 Frascati, Roma (Italy); Mazon, Didier [Association EURATOM-CEA, CEA Cadarache, DSM/IRFM, 13108 St. Paul Lez Durance Cedex (France)

    2013-08-21

    A triple gas electron multiplier (GEM) detector has been built and characterized in a collaboration between ENEA, INFN and CEA to develop a soft X-ray imaging diagnostic for magnetic fusion plasmas. It has an active area of 5×5 cm{sup 2}, 128 pixels and electronics in counting mode. Since burning plasma experiments will have a very large background of radiation, this prototype has been tested with contemporary X-ray, neutron and gamma irradiation, to study the detection efficiencies, and the discrimination capabilities. The detector has been preliminarily characterized under DD neutron irradiation (2.45 MeV) up to 2.2×10{sup 6} n/s on the detector active area, showing a detection efficiency of about 10{sup −4}, while the detection efficiency of X-rays is more than three orders of magnitude higher. The detector has been also tested under DT neutron flux (14 MeV) up to 2.8×10{sup 8} n/s on the whole detector, with a detection efficiency of about 10{sup −5}. The calibration of the γ-rays detection has been done by means of a source of {sup 60}Co (gamma rays of energy 1.17 MeV and 1.33 MeV) and the detection efficiency was found of the order of 10{sup −4}. Thanks to the adjustable gain of the detector and the discrimination threshold of the electronics, it is possible to minimize the sensitivity to neutrons and gamma, and discriminate the X-ray signals even with very high radiative background.

  11. A comparative analysis of OTF, NPS, and DQE in energy integrating and photon counting digital x-ray detectors

    International Nuclear Information System (INIS)

    Acciavatti, Raymond J.; Maidment, Andrew D. A.

    2010-01-01

    Purpose: One of the benefits of photon counting (PC) detectors over energy integrating (EI) detectors is the absence of many additive noise sources, such as electronic noise and secondary quantum noise. The purpose of this work is to demonstrate that thresholding voltage gains to detect individual x rays actually generates an unexpected source of white noise in photon counters. Methods: To distinguish the two detector types, their point spread function (PSF) is interpreted differently. The PSF of the energy integrating detector is treated as a weighting function for counting x rays, while the PSF of the photon counting detector is interpreted as a probability. Although this model ignores some subtleties of real imaging systems, such as scatter and the energy-dependent amplification of secondary quanta in indirect-converting detectors, it is useful for demonstrating fundamental differences between the two detector types. From first principles, the optical transfer function (OTF) is calculated as the continuous Fourier transform of the PSF, the noise power spectra (NPS) is determined by the discrete space Fourier transform (DSFT) of the autocovariance of signal intensity, and the detective quantum efficiency (DQE) is found from combined knowledge of the OTF and NPS. To illustrate the calculation of the transfer functions, the PSF is modeled as the convolution of a Gaussian with the product of rect functions. The Gaussian reflects the blurring of the x-ray converter, while the rect functions model the sampling of the detector. Results: The transfer functions are first calculated assuming outside noise sources such as electronic noise and secondary quantum noise are negligible. It is demonstrated that while OTF is the same for two detector types possessing an equivalent PSF, a frequency-independent (i.e., ''white'') difference in their NPS exists such that NPS PC ≥NPS EI and hence DQE PC ≤DQE EI . The necessary and sufficient condition for equality is that the PSF

  12. MHz rate X-Ray imaging with GaAs:Cr sensors using the LPD detector system

    Science.gov (United States)

    Veale, M. C.; Booker, P.; Cline, B.; Coughlan, J.; Hart, M.; Nicholls, T.; Schneider, A.; Seller, P.; Pape, I.; Sawhney, K.; Lozinskaya, A. D.; Novikov, V. A.; Tolbanov, O. P.; Tyazhev, A.; Zarubin, A. N.

    2017-02-01

    The STFC Rutherford Appleton Laboratory (U.K.) and Tomsk State University (Russia) have been working together to develop and characterise detector systems based on chromium-compensated gallium arsenide (GaAs:Cr) semiconductor material for high frame rate X-ray imaging. Previous work has demonstrated the spectroscopic performance of the material and its resistance to damage induced by high fluxes of X-rays. In this paper, recent results from experiments at the Diamond Light Source Synchrotron have demonstrated X-ray imaging with GaAs:Cr sensors at a frame rate of 3.7 MHz using the Large Pixel Detector (LPD) ASIC, developed by STFC for the European XFEL. Measurements have been made using a monochromatic 20 keV X-ray beam delivered in a single hybrid pulse with an instantenous flux of up to ~ 1 × 1010 photons s-1 mm-2. The response of 500 μm GaAs:Cr sensors is compared to that of the standard 500 μm thick LPD Si sensors.

  13. X-ray imaging with sub-micron resolution using large-area photon counting detectors Timepix

    Science.gov (United States)

    Dudak, J.; Karch, J.; Holcova, K.; Zemlicka, J.

    2017-12-01

    As X-ray micro-CT became a popular tool for scientific purposes a number of commercially available CT systems have emerged on the market. Micro-CT systems have, therefore, become widely accessible and the number of research laboratories using them constantly increases. However, even when CT scans with spatial resolution of several micrometers can be performed routinely, data acquisition with sub-micron precision remains a complicated task. Issues come mostly from prolongation of the scan time inevitably connected with the use of nano-focus X-ray sources. Long exposure time increases the noise level in the CT projections. Furthermore, considering the sub-micron resolution even effects like source-spot drift, rotation stage wobble or thermal expansion become significant and can negatively affect the data. The use of dark-current free photon counting detectors as X-ray cameras for such applications can limit the issue of increased image noise in the data, however the mechanical stability of the whole system still remains a problem and has to be considered. In this work we evaluate the performance of a micro-CT system equipped with nano-focus X-ray tube and a large area photon counting detector Timepix for scans with effective pixel size bellow one micrometer.

  14. Development of a novel direct X-ray detector using photoinduced discharge (PID) readout for digital radiography

    Science.gov (United States)

    Heo, D.; Jeon, S.; Kim, J.-S.; Kim, R. K.; Cha, B. K.; Moon, B. J.; Yoon, J.

    2013-02-01

    We developed a novel direct X-ray detector using photoinduced discharge (PID) readout for digital radiography. The pixel resolution is 512 × 512 with 200 μm pixel and the overall active dimensions of the X-ray imaging panel is 10.24 cm × 10.24 cm. The detector consists of an X-ray absorption layer of amorphous selenium, a charge accumulation layer of metal, and a PID readout layer of amorphous silicon. In particular, the charge accumulation is pixelated because image charges generated by X-ray should be stored pixel by pixel. Here the image charges, or holes, are recombined with electrons generated by the PID method. We used a 405 nm laser diode and cylindrical lens to make a line beam source with a width of 50 μm for PID readout, which generates charges for each pixel lines during the scan. We obtained spatial frequencies of about 1.0 lp/mm for the X-direction (lateral direction) and 0.9 lp/mm for the Y-direction (scanning direction) at 50% modulation transfer function.

  15. SERENDIPITOUS DETECTION OF X-RAY EMISSION FROM THE HOT BORN-AGAIN CENTRAL STAR OF THE PLANETARY NEBULA K 1-16

    Energy Technology Data Exchange (ETDEWEB)

    Montez, Rodolfo Jr. [Department of Physics and Astronomy, Vanderbilt University, Nashville, TN 37235 (United States); Kastner, Joel H., E-mail: rodolfo.montez.jr@gmail.com, E-mail: jhk@cis.rit.edu [Center for Imaging Science and Laboratory for Multiwavelength Astrophysics, Rochester Institute of Technology, 54 Lomb Memorial Drive, Rochester, NY 14623 (United States)

    2013-03-20

    We report the serendipitous detection of point-like X-ray emission from the hot, PG1159-type central star of the planetary nebula (CSPN) K 1-16 by the XMM-Newton and Chandra X-Ray Observatories. The CSPN lies superimposed on a galaxy cluster that includes an X-ray-bright quasar, but we have successfully isolated the CSPN X-ray emission from the strong diffuse background contributed by the quasar and intracluster gas. We have modeled the XMM-Newton and Chandra X-ray data, taking advantage of the contrasting detection efficiencies of the two observatories to better constrain the low-energy spectral response of Chandra's Advanced CCD Imaging Spectrometer. We find that the CSPN X-ray spectrum is well characterized by the combination of a non-local thermodynamic equilibrium model atmosphere with T{sub *} {approx} 135 kK and a carbon-rich, optically thin thermal plasma with T{sub X} {approx} 1 MK. These results for X-ray emission from the K 1-16 CSPN, combined with those obtained for other PG1159-type objects, lend support to the 'born-again' scenario for Wolf-Rayet and PG1159 CSPNe, wherein a late helium shell flash dredges up carbon-rich intershell material and ejects this material into the circumstellar environment.

  16. A gas microstrip detector for X-ray imaging with readout of the anode by resistive division

    CERN Document Server

    Bateman, J E; Lodge, A B; Stephenson, R; Mutikainen, R; Suni, I; Morse, J

    2002-01-01

    The results are presented of a development programme aimed at the validation of the key concepts and technologies for the construction of a two-dimensional X-ray detector based on gas microstrip detector technology using resistive division along the anode to achieve the second dimension. A prototype detector and its associated electronic readout system have been developed which demonstrate the capability of a spatial resolution (standard deviation) of approximately ((1)/(1000)) of the working aperture combined with readout rates of up to 400 kHz per anode. Test results and a description of the position sensing circuitry are given.

  17. CdTe and CdZnTe detectors behavior in X-ray computed tomography conditions

    CERN Document Server

    Ricq, S; Garcin, M

    2000-01-01

    The application of CdTe and CdZnTe 2D array detectors for medical X-ray Computed Tomography (XCT) is investigated. Different metallic electrodes have been deposited on High-Pressure Bridgman Method CdZnTe and on Traveling Heater Method CdTe:Cl. These detectors are exposed to X-rays in the CT irradiation conditions and are characterized experimentally in current mode. Detectors performances such as sensitivity and response speed are studied. They are correlated with charge trapping and de-trapping. The trapped carrier space charges may influence the injection from the electrodes. This enables one to get information on the nature of the predominant levels involved. The performances achieved are encouraging: dynamic ranges higher than 4 decades and current decreases of 3 decades in 4 ms after X-ray beam cut-off are obtained. Nevertheless, these detectors are still limited by high trap densities responsible for the memory effect that makes them unsuitable for XCT.

  18. Monte Carlo analysis of megavoltage x-ray interaction-induced signal and noise in detectors for container inspection

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Jinwoo; Park, Jiwoong; Kim, Junwoo; Kim, Dong Woon; Kim, Ho Kyung [Pusan National University, Busan (Korea, Republic of); Lim, Chang Hwy [Korea Research Institute of Ships and Ocean Engineering, Daejeon (Korea, Republic of)

    2016-10-15

    In a scanner system, a scintillation crystal is the first stage in the cascaded imaging chain transferring x-ray interaction information in cargo to be investigated to the final user who investigates x-ray images. On the other hand, the signal and noise is irreversibly transferred through the cascaded imaging chain. Therefore, the imaging performance of the first stage scintillator mainly governs the ultimate imaging performance of the system. In MV imaging, it is generally accepted that high-density scintillators, because of their sufficient optical yield, and low optical self-absorption and scattering coefficients. We chose the CdWO{sub 4} as the scintillation material. CdWO{sub 4} has a high density (7.9 g/cm{sup 3}), high atomic number (64), resistance to radiation, high optical yield, and low optical self-absorption. For the given MV spectrum, the improvement of QE from a detector with a thickness of 10 mm to 30 mm is 27% whereas the improvement from 30 mm to 50 mm is only 7%. On the other hand, the Swank noise is almost independent of the detector thickness. Consequently, the improvement of DQE from a detector with a thickness of 10 mm to 30 mm is 46% whereas the improvement from 30 mm to 50 mm is only 11%. In conclusion, the detector thickness of 30 mm would be the best for x-ray interaction-induced signal and noise performance as well as cost.

  19. A unified statistical framework for material decomposition using multienergy photon counting x-ray detectors

    International Nuclear Information System (INIS)

    Choi, Jiyoung; Kang, Dong-Goo; Kang, Sunghoon; Sung, Younghun; Ye, Jong Chul

    2013-01-01

    Purpose: Material decomposition using multienergy photon counting x-ray detectors (PCXD) has been an active research area over the past few years. Even with some success, the problem of optimal energy selection and three material decomposition including malignant tissue is still on going research topic, and more systematic studies are required. This paper aims to address this in a unified statistical framework in a mammographic environment.Methods: A unified statistical framework for energy level optimization and decomposition of three materials is proposed. In particular, an energy level optimization algorithm is derived using the theory of the minimum variance unbiased estimator, and an iterative algorithm is proposed for material composition as well as system parameter estimation under the unified statistical estimation framework. To verify the performance of the proposed algorithm, the authors performed simulation studies as well as real experiments using physical breast phantom and ex vivo breast specimen. Quantitative comparisons using various performance measures were conducted, and qualitative performance evaluations for ex vivo breast specimen were also performed by comparing the ground-truth malignant tissue areas identified by radiologists.Results: Both simulation and real experiments confirmed that the optimized energy bins by the proposed method allow better material decomposition quality. Moreover, for the specimen thickness estimation errors up to 2 mm, the proposed method provides good reconstruction results in both simulation and real ex vivo breast phantom experiments compared to existing methods.Conclusions: The proposed statistical framework of PCXD has been successfully applied for the energy optimization and decomposition of three material in a mammographic environment. Experimental results using the physical breast phantom and ex vivo specimen support the practicality of the proposed algorithm

  20. Efficiency calibration of x-ray HPGe detectors for photons with energies above the Ge K binding energy

    Energy Technology Data Exchange (ETDEWEB)

    Maidana, Nora L., E-mail: nmaidana@if.usp.br [Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP:05508-900 São Paulo, SP (Brazil); Vanin, Vito R.; Jahnke, Viktor [Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP:05508-900 São Paulo, SP (Brazil); Fernández-Varea, José M. [Facultat de Física (ECM and ICC), Universitat de Barcelona, Diagonal 645, E-08028 Barcelona (Spain); Martins, Marcos N. [Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP:05508-900 São Paulo, SP (Brazil); Brualla, Lorenzo [NCTeam, Strahlenklinik, Universitätsklinikum Essen, Hufelandstraße 55, D-45122 Essen (Germany)

    2013-11-21

    We report on the efficiency calibration of a HPGe x-ray detector using radioactive sources and an analytical expression taken from the literature, in two different arrangements, with and without a broad-angle collimator. The frontal surface of the Ge crystal was scanned with pencil beams of photons. The Ge dead layer was found to be nonuniform, with central and intermediate regions that have thin (μm range) and thick (mm range) dead layers, respectively, surrounded by an insensitive ring. We discuss how this fact explains the observed efficiency curves and generalize the adopted model. We show that changes in the thickness of the Ge-crystal dead layer affect the efficiency of x-ray detectors, but the use of an appropriate broad-beam external collimator limiting the photon flux to the thin dead layer in the central region leads to the expected efficiency dependence with energy and renders the calibration simpler.

  1. Development of flat panel X-ray detector utilizing a CdZnTe film as conversion layer

    International Nuclear Information System (INIS)

    Tokuda, Satoshi; Kishihara, Hiroyuki; Kaino, Masatomo; Sato, Toshiyuki

    2006-01-01

    A polycrystalline CdZnTe film formed by the CSS (closed-spaced sublimation) method is one of the most promising materials as a conversion layer of next-generation highly efficient flat-panel X-ray detectors. Therefore, we have developed a prototype of a new flat-panel X-ray detector (a sensing region of 3 inches by 3 inches) with the film and evaluated its commercial feasibility. This paper describes evaluation of the physical and imaging properties of the prototype and explains the features of the CdZnTe film and the construction, specifications, and fabrication procedures of the prototype. Also included in this paper are formation of a semiconductor thin film barrier layer by the CBD (chemical bath deposition) method and conjunction of a sensor substrate and a TFT array substrate with the bump electrodes formed by screen printing, both of which we have developed during the course of the development of the prototype. (author)

  2. Design and theoretical investigation of a digital x-ray detector with large area and high spatial resolution

    Science.gov (United States)

    Gui, Jianbao; Guo, Jinchuan; Yang, Qinlao; Liu, Xin; Niu, Hanben

    2007-05-01

    X-ray phase contrast imaging is a promising new technology today, but the requirements of a digital detector with large area, high spatial resolution and high sensitivity bring forward a large challenge to researchers. This paper is related to the design and theoretical investigation of an x-ray direct conversion digital detector based on mercuric iodide photoconductive layer with the latent charge image readout by photoinduced discharge (PID). Mercuric iodide has been verified having a good imaging performance (high sensitivity, low dark current, low voltage operation and good lag characteristics) compared with the other competitive materials (α-Se,PbI II,CdTe,CdZnTe) and can be easily deposited on large substrates in the manner of polycrystalline. By use of line scanning laser beam and parallel multi-electrode readout make the system have high spatial resolution and fast readout speed suitable for instant general radiography and even rapid sequence radiography.

  3. Correction of diagnostic x-ray spectra measured with CdTe and CdZnTe detectors

    Energy Technology Data Exchange (ETDEWEB)

    Matsumoto, M [Osaka Univ., Suita (Japan). Medical School; Kanamori, H; Toragaito, T; Taniguchi, A

    1996-07-01

    We modified the formula of stripping procedure presented by E. Di. Castor et al. We added the Compton scattering and separated K{sub {alpha}} radiation of Cd and Te (23 and 27keV, respectively). Using the new stripping procedure diagnostic x-ray spectra (object 4mm-Al) of tube voltage 50kV to 100kV for CdTe and CdZnTe detectors are corrected with comparison of those spectra for the Ge detector. The corrected spectra for CdTe and CdZnTe detectors coincide with those for Ge detector at lower tube voltage than 70kV. But the corrected spectra at higher tube voltage than 70kV do not coincide with those for Ge detector. The reason is incomplete correction for full energy peak efficiencies of real CdTe and CdZnTe detectors. (J.P.N.)

  4. The PASERO Project: parallel and serial readout systems for gas proportional synchrotron radiation X-ray detectors

    CERN Document Server

    Koch, M H J; Briquet-Laugier, F; Epstein, A; Sheldon, S; Beloeuvre, E; Gabriel, A; Hervé, C; Kocsis, M; Koschuch, A; Laggner, P; Leingartner, W; Raad-Iseli, C D; Reimann, T; Golding, F; Torki, K

    2001-01-01

    A project aiming at producing more efficient position sensitive gas proportional detectors and readout systems is presented. An area detector with reduced electrode spacing and a spatial resolution of 0.5 mm and two time to digital convertors (TDC) based on ASICs were produced. The first TDC, intended for use with linear detectors, relies on time to space conversion, whereas the second one, for area detectors, uses a ring oscillator with a phase locked loop. A parallel readout system for multi-anode detectors aiming at a maximum count rate extensively uses RISC microcontrollers. An electronic simulator of linear detectors built for test purposes and a mechanical chopper used for attenuation of the X-ray beam are also briefly described.

  5. Upsurge of X-ray astronomy 230-

    International Nuclear Information System (INIS)

    Hudec, D.R.

    1978-01-01

    Instruments are described used for X-ray astronomy, namely X-ray detectors and X-ray telescopes. Unlike telescopes, the detectors do not comprise X-ray optics. A survey is given of the results obtained in solar and stellar X-ray astronomy and hypotheses are submitted on the origin of X radiation in the interstellar space. (J.B.)

  6. Mathematical Formalism for Designing Wide-Field X-Ray Telescopes: Mirror Nodal Positions and Detector Tilts

    Science.gov (United States)

    Elsner, R. F.; O'Dell, S. L.; Ramsey, B. D.; Weisskopf, M. C.

    2011-01-01

    We provide a mathematical formalism for optimizing the mirror nodal positions along the optical axis and the tilt of a commonly employed detector configuration at the focus of a x-ray telescope consisting of nested mirror shells with known mirror surface prescriptions. We adopt the spatial resolution averaged over the field-of-view as the figure of merit M. A more complete description appears in our paper in these proceedings.

  7. In-flight performance of the soft x-ray spectrometer detector system on Astro-H

    Science.gov (United States)

    Porter, Frederick S.; Boyce, Kevin R.; Chiao, Meng P.; Eckart, Megan E.; Fujimoto, Ryuichi; Ishisaki, Yoshitaka; Kilbourne, Caroline Anne; Leutenegger, Maurice A.; McCammon, Daniel; Mitsuda, Kazuhisa; Sato, Kosuke; Seta, Hiromi; Sawada, Makoto; Sneiderman, Gary A.; Szymkowiak, Andrew E.; Takei, Yoh; Tashiro, Makoto S.; Tsujimoto, Masahiro; Watanabe, Tomomi; Yamada, Shinya

    2018-01-01

    The soft x-ray spectrometer (SXS) instrument was launched aboard the Astro-H (Hitomi) observatory on February 17, 2016. The SXS is based on a high-sensitivity x-ray calorimeter detector system that has been successfully deployed in many ground and suborbital spectrometers. The instrument was to provide essential diagnostics for nearly every class of x-ray emitting objects from the atmosphere of Jupiter to the outskirts of galaxy clusters, without degradation for spatially extended objects. The SXS detector system consisted of a 36-pixel cryogenic microcalorimeter array operated at a heat sink temperature of 50 mK. In preflight testing, the detector system demonstrated a resolving power of better than 1300 at 6 keV with a simultaneous bandpass from below 0.3 keV to above 12 keV with a timing precision better than 100 μs. In addition, a solid-state anticoincidence detector was placed directly behind the detector array for background suppression. The detector error budget included the measured interference from the SXS cooling system and the spacecraft. Additional margin for on-orbit gain stability and on-orbit spacecraft interference were also included predicting an on-orbit performance that meets or exceeds the 7-eV FWHM at 6-keV requirement. The actual on-orbit spectral resolution was better than 5 eV FWHM at 6 keV, easily satisfying the instrument requirement. Here, we discuss the actual on-orbit performance of the SXS detector system and compare this to performance in preflight testing and the on-orbit predictions. We will also discuss the on-orbit gain stability, additional on-orbit interference, and measurements of the on-orbit background.

  8. Semi-empirical procedures for correcting detector size effect on clinical MV x-ray beam profiles

    International Nuclear Information System (INIS)

    Sahoo, Narayan; Kazi, Abdul M.; Hoffman, Mark

    2008-01-01

    The measured radiation beam profiles need to be corrected for the detector size effect to derive the real profiles. This paper describes two new semi-empirical procedures to determine the real profiles of high-energy x-ray beams by removing the detector size effect from the measured profiles. Measured profiles are corrected by shifting the position of each measurement point by a specific amount determined from available theoretical and experimental knowledge in the literature. The authors developed two procedures to determine the amount of shift. In the first procedure, which employs the published analytical deconvolution procedure of other investigators, the shift is determined from the comparison of the analytical fit of the measured profile and the corresponding analytical real profile derived from the deconvolution of the fitted measured profile and the Gaussian detector response function. In the second procedure, the amount of shift at any measurement point is considered to be proportional to the value of an analytical function related to the second derivative of the real profile at that point. The constant of proportionality and a parameter in the function are obtained from the values of the shifts at the 90%, 80%, 20%, and 10% dose levels, which are experimentally known from the published results of other investigators to be approximately equal to half of the radius of the detector. These procedures were tested by correcting the profiles of 6 and 18 MV x-ray beams measured by three different ionization chambers and a stereotactic field diode detector with 2.75, 2, 1, and 0.3 mm radii of their respective active cylindrical volumes. The corrected profiles measured by different detectors are found to be in close agreement. The detector size corrected penumbra widths also agree with the expected values based on the results of an earlier investigation. Thus, the authors concluded that the proposed procedures are accurate and can be used to derive the real

  9. Review of medical imaging with emphasis on X-ray detectors

    Science.gov (United States)

    Hoheisel, Martin

    2006-07-01

    Medical imaging can be looked at from two different perspectives, the medical and the physical. The medical point of view is application-driven and involves finding the best way of tackling a medical problem through imaging, i.e. either to answer a diagnostic question, or to facilitate a therapy. For this purpose, industry offers a broad spectrum of radiographic, fluoroscopic, and angiographic equipment. The requirements depend on the medical problem: which organs have to be imaged, which details have to be made visible, how to deal with the problem of motion if any, and so forth. In radiography, for instance, large detector sizes of up to 43 cm×43 cm and relatively high energies are needed to image a whole chest. In mammography, pixel sizes between 25 and 70 μm are favorable for good spatial resolution, which is essential for detecting microcalcifications. In cardiology, 30-60 images per second are required to follow the heart's motion. In computed tomography, marginal contrast differences down to one Hounsfield unit have to be resolved. In all cases, but especially in pediatrics, the required radiation dose must be kept as low as reasonably achievable. Moreover, three-dimensional(3D) reconstruction of image data allows much better orientation in the body, permitting a more accurate diagnosis, precise treatment planning, and image-guided therapy. Additional functional information from different modalities is very helpful, information such as perfusion, flow rate, diffusion, oxygen concentration, metabolism, and receptor affinity for specific molecules. To visualize, functional and anatomical information are fused into one combined image. The physical point of view is technology-driven. A choice of different energies from the electromagnetic spectrum is available for imaging; not only X-rays in the range of 10-150 keV, but also γ rays, which are used in nuclear medicine, X-rays in the MeV range, which are used in portal imaging to monitor radiation therapy

  10. Review of medical imaging with emphasis on X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Hoheisel, Martin [Siemens AG Medical Solutions, Angiography, Fluoroscopic- and Radiographic Systems, Innovations, Siemensstr.1, 91301 Forchheim (Germany)]. E-mail: martin.hoheisel@siemens.com

    2006-07-01

    Medical imaging can be looked at from two different perspectives, the medical and the physical. The medical point of view is application-driven and involves finding the best way of tackling a medical problem through imaging, i.e. either to answer a diagnostic question, or to facilitate a therapy. For this purpose, industry offers a broad spectrum of radiographic, fluoroscopic, and angiographic equipment. The requirements depend on the medical problem: which organs have to be imaged, which details have to be made visible, how to deal with the problem of motion if any, and so forth. In radiography, for instance, large detector sizes of up to 43 cmx43 cm and relatively high energies are needed to image a whole chest. In mammography, pixel sizes between 25 and 70 {mu}m are favorable for good spatial resolution, which is essential for detecting microcalcifications. In cardiology, 30-60 images per second are required to follow the heart's motion. In computed tomography, marginal contrast differences down to one Hounsfield unit have to be resolved. In all cases, but especially in pediatrics, the required radiation dose must be kept as low as reasonably achievable. Moreover, three-dimensional(3D) reconstruction of image data allows much better orientation in the body, permitting a more accurate diagnosis, precise treatment planning, and image-guided therapy. Additional functional information from different modalities is very helpful, information such as perfusion, flow rate, diffusion, oxygen concentration, metabolism, and receptor affinity for specific molecules. To visualize, functional and anatomical information are fused into one combined image. The physical point of view is technology-driven. A choice of different energies from the electromagnetic spectrum is available for imaging; not only X-rays in the range of 10-150 keV, but also {gamma} rays, which are used in nuclear medicine, X-rays in the MeV range, which are used in portal imaging to monitor radiation

  11. Review of medical imaging with emphasis on X-ray detectors

    International Nuclear Information System (INIS)

    Hoheisel, Martin

    2006-01-01

    Medical imaging can be looked at from two different perspectives, the medical and the physical. The medical point of view is application-driven and involves finding the best way of tackling a medical problem through imaging, i.e. either to answer a diagnostic question, or to facilitate a therapy. For this purpose, industry offers a broad spectrum of radiographic, fluoroscopic, and angiographic equipment. The requirements depend on the medical problem: which organs have to be imaged, which details have to be made visible, how to deal with the problem of motion if any, and so forth. In radiography, for instance, large detector sizes of up to 43 cmx43 cm and relatively high energies are needed to image a whole chest. In mammography, pixel sizes between 25 and 70 μm are favorable for good spatial resolution, which is essential for detecting microcalcifications. In cardiology, 30-60 images per second are required to follow the heart's motion. In computed tomography, marginal contrast differences down to one Hounsfield unit have to be resolved. In all cases, but especially in pediatrics, the required radiation dose must be kept as low as reasonably achievable. Moreover, three-dimensional(3D) reconstruction of image data allows much better orientation in the body, permitting a more accurate diagnosis, precise treatment planning, and image-guided therapy. Additional functional information from different modalities is very helpful, information such as perfusion, flow rate, diffusion, oxygen concentration, metabolism, and receptor affinity for specific molecules. To visualize, functional and anatomical information are fused into one combined image. The physical point of view is technology-driven. A choice of different energies from the electromagnetic spectrum is available for imaging; not only X-rays in the range of 10-150 keV, but also γ rays, which are used in nuclear medicine, X-rays in the MeV range, which are used in portal imaging to monitor radiation therapy

  12. Monte Carlo simulation of the X-ray response of a germanium microstrip detector with energy and position resolution

    CERN Document Server

    Rossi, G; Fajardo, P; Morse, J

    1999-01-01

    We present Monte Carlo computer simulations of the X-ray response of a micro-strip germanium detector over the energy range 30-100 keV. The detector consists of a linear array of lithographically defined 150 mu m wide strips on a high purity monolithic germanium crystal of 6 mm thickness. The simulation code is divided into two parts. We first consider a 10 mu m wide X-ray beam striking the detector surface at normal incidence and compute the interaction processes possible for each photon. Photon scattering and absorption inside the detector crystal are simulated using the EGS4 code with the LSCAT extension for low energies. A history of events is created of the deposited energies which is read by the second part of the code which computes the energy histogram for each detector strip. Appropriate algorithms are introduced to account for lateral charge spreading occurring during charge carrier drift to the detector surface, and Fano and preamplifier electronic noise contributions. Computed spectra for differen...

  13. Investigation of the radiation leakage from X ray flaw detectors and the improvement measures for the unqualified products

    International Nuclear Information System (INIS)

    Li Yiachun; Wu Yi; Pang Hu; Bai Bin

    1997-01-01

    The authors introduce investigation methods and results for radiation leakage from X ray flaw detectors, which are used in Beijing area. Total 21 sets of flaw detectors made in 8 factories in Beijing, Shanghai etc. have been tested, of which 16 sets made in Beijing, Dandong and Japan are gas cooling flaw detectors, and rest 5 sets made in Shanghai and Germany are water or oil cooling detectors. The air Kerma rate of leakage radiation at 1 m from the X ray tube target were measured by Type FJ-347A X, γ dosimeter. It can be seen from the results that, compared with the trade standard ZBY315-83, 5 sets of water or oil cooling flaw detectors are all qualified. However, only two sets of gas cooling detectors are qualified, and the radiation leakage of another 14 sets are over the values specified in the standard. The reason is analyzed, and some advices about the measures of improving radiation protection structure design and production technology for the unqualified products have been proposed

  14. High-spatial resolution and high-spectral resolution detector for use in the measurement of solar flare hard x rays

    International Nuclear Information System (INIS)

    Desai, U.D.; Orwig, L.E.

    1988-01-01

    In the areas of high spatial resolution, the evaluation of a hard X-ray detector with 65 micron spatial resolution for operation in the energy range from 30 to 400 keV is proposed. The basic detector is a thick large-area scintillator faceplate, composed of a matrix of high-density scintillating glass fibers, attached to a proximity type image intensifier tube with a resistive-anode digital readout system. Such a detector, combined with a coded-aperture mask, would be ideal for use as a modest-sized hard X-ray imaging instrument up to X-ray energies as high as several hundred keV. As an integral part of this study it was also proposed that several techniques be critically evaluated for X-ray image coding which could be used with this detector. In the area of high spectral resolution, it is proposed to evaluate two different types of detectors for use as X-ray spectrometers for solar flares: planar silicon detectors and high-purity germanium detectors (HPGe). Instruments utilizing these high-spatial-resolution detectors for hard X-ray imaging measurements from 30 to 400 keV and high-spectral-resolution detectors for measurements over a similar energy range would be ideally suited for making crucial solar flare observations during the upcoming maximum in the solar cycle

  15. Microchannel plates as detectors and amplifiers of x-ray images

    International Nuclear Information System (INIS)

    Wiedwald, J.D.

    1992-08-01

    Two decades of development driven largely by military night vision applications has led to the availability of a wide selection of microchannel plates for use by the scientific community. Microchannel plates (MCPs) are electron multipliers which retain a high degree of spatial resolution making it possible to amplify electron images by factors of 1,000 or more. Plates having 40 mm diameter and intrinsic spatial resolution of 8 μm are readily available. By coating the front surface of a microchannel plate with an x-ray sensitive photocathode material, x-ray images can be detected and amplified. While the detective quantum efficiency is relatively low, the low noise of the MCP (including the ability to construct images by single photon detection) and its high dynamic range make it suitable for some x-ray microscopy applications. The principles of MCP operation and typical performance are discussed. Examples of related applications and commercial capabilities are also presented

  16. Phonon-mediated superconducting transition-edge sensor X-ray detectors for use in astronomy

    Science.gov (United States)

    Leman, Steven W.; Martinez-Galarce, Dennis S.; Brink, Paul L.; Cabrera, Blas; Castle, Joseph P.; Morse, Kathleen; Stern, Robert A.; Tomada, Astrid

    2004-09-01

    Superconducting Transition-Edge Sensors (TESs) are generating a great deal of interest in the areas of x-ray astrophysics and space science, particularly to develop them as large-array, imaging x-ray spectrometers. We are developing a novel concept that is based on position-sensitive macro-pixels placing TESs on the backside of a silicon or germanium absorber. Each x-ray absorbed will be position (X/δX and Y/δY ~ 100) and energy (E/δE ~ 1000) resolved via four distributed TES readouts. In the future, combining such macropixels with advances in multiplexing could lead to 30 by 30 arrays of close-packed macro-pixels equivalent to imaging instruments of 10 megapixels or more. We report on our progress to date and discuss its application to a plausible solar satellite mission and plans for future development.

  17. The imaging pin detector - a simple and effective new imaging device for soft x-rays and soft beta emissions

    International Nuclear Information System (INIS)

    Bateman, J.E.

    1984-01-01

    The development of a new bidimensional imaging detector system for soft X and beta radiations is reported. Based on the detection of the differential induction signals on pickup electrodes placed around a point anode in a gas avalanche detector, the system described has achieved a spatial resolution of better than 1mm fwhm over a field of 30mm diameter while preserving excellent pulse height resolution. The present device offers considerable potential as a cheap and robust imaging system for applications in X-ray diffraction and autoradiography. (author)

  18. Phase contrast imaging: Effect of increased object-detector distances at X-ray diagnostic and megavoltage energies

    Energy Technology Data Exchange (ETDEWEB)

    Loveland, J.; Gundogdu, O. [Department of Physics, University of Surrey, Guildford, Surrey GU2 7XH (United Kingdom); Morton, E. [Rapiscan Systems, Units 2,3,4, Radnor Park Trading Estate, Congleton, Cheshire CW12 4XJ (United Kingdom); Wells, K. [CVSSP, University of Surrey, Guildford, Surrey GU2 7XH (United Kingdom); Bradley, D.A., E-mail: d.a.bradley@surrey.ac.uk [Department of Physics, University of Surrey, Guildford, Surrey GU2 7XH (United Kingdom)

    2011-10-01

    The effect of varying object to detector separation at constant and varying magnification has been investigated at an accelerating potential of 30 kVp. Edge-contrast enhancement provided by phase effects was investigated for a drinking straw and found to provide up to 2.52{+-}0.02x the contrast for a PVC Heaviside step function. An optimum magnification of 1.5x was found to apply for the microfocus X-ray tube setup used. Imaging at nominal megavoltage energies was investigated using a Rapiscan Systems Eagle M4500 series scanner. For a fixed source-detector separation, increased magnification improved edge contrast and spatial resolution.

  19. Phase contrast imaging: Effect of increased object-detector distances at X-ray diagnostic and megavoltage energies

    International Nuclear Information System (INIS)

    Loveland, J.; Gundogdu, O.; Morton, E.; Wells, K.; Bradley, D.A.

    2011-01-01

    The effect of varying object to detector separation at constant and varying magnification has been investigated at an accelerating potential of 30 kVp. Edge-contrast enhancement provided by phase effects was investigated for a drinking straw and found to provide up to 2.52±0.02x the contrast for a PVC Heaviside step function. An optimum magnification of 1.5x was found to apply for the microfocus X-ray tube setup used. Imaging at nominal megavoltage energies was investigated using a Rapiscan Systems Eagle M4500 series scanner. For a fixed source-detector separation, increased magnification improved edge contrast and spatial resolution.

  20. X-ray detectors for diffraction studies and their use with synchrotron radiation

    International Nuclear Information System (INIS)

    Milch, J.

    1976-02-01

    All techniques for X-ray diffraction studies on biological materials exhibit certain limitations. The characteristics of several X-ray detection systems, namely film, multiwire proportional counter and image intensified TV, are discussed and compared for application to specific biological studies. For the high count-rate situation existing at a synchrotron, it is shown that film is a good choice, but that the image intensified TV exhibits significant advantages. The details of such a system now being used at Princeton with a low intensity source are given and current results presented

  1. Discriminated neutron and X-ray radiography using multi-color scintillation detector

    International Nuclear Information System (INIS)

    Nittoh, Koichi; Takahara, Takeshi; Yoshida, Tadashi; Tamura, Toshiyuki

    1999-01-01

    A new conversion screen Gd 2 O 2 S:Eu is developed, which emits red light on irradiation by thermal neutrons. By applying this in combination with the currently used Gd 2 O 2 S:Tb, a green-light scintillator, in the radiography under a neutron + X-ray coexisting field, we can easily separate the neutron image and the X-ray image by simple color-image processing. This technique enables a non-destructive and detailed inspection of industrial products composed both of light elements (water, plastics, etc.) and heavy elements (metals), widening the horizon of new applications

  2. Discriminated neutron and X-ray radiography using multi-color scintillation detector

    CERN Document Server

    Nittoh, K; Yoshida, T; Tamura, T

    1999-01-01

    A new conversion screen Gd sub 2 O sub 2 S:Eu is developed, which emits red light on irradiation by thermal neutrons. By applying this in combination with the currently used Gd sub 2 O sub 2 S:Tb, a green-light scintillator, in the radiography under a neutron + X-ray coexisting field, we can easily separate the neutron image and the X-ray image by simple color-image processing. This technique enables a non-destructive and detailed inspection of industrial products composed both of light elements (water, plastics, etc.) and heavy elements (metals), widening the horizon of new applications.

  3. A High Position Resolution X-ray Detector: an Edge on Illuminated Capillary Plate Combined with a Gas Amplification Structure

    CERN Document Server

    Iacobaeus, C.; Lund-Jensen, B.; Ostling, J.; Pavlopoulos, P.; Peskov, V.; Tokanai, F.

    2006-01-01

    We have developed and successfully tested a prototype of a new type of high position resolution hybrid X-ray detector. It contains a thin wall lead glass capillary plate converter of X-rays combined with a microgap parallel-plate avalanche chamber filled with gas at 1 atm. The operation of these converters was studied in a wide range of X-ray energies (from 6 to 60 keV) at incident angles varying from 0-90 degree. The detection efficiency, depending on the geometry, photon energy, incident angle and the mode of operation, was between 5-30 percent in a single step mode and up to 50 percent in a multi-layered combination. Depending on the capillary geometry, the position resolution achieved was between 0.050-0.250 mm in digital form and was practically independent of the photon energy or gas mixture. The usual lead glass capillary plates operated without noticeable charging up effects at counting rates of 50 Hz/mm2, and hydrogen treated capillaries up to 10E5 Hz/mm2. The developed detector may open new possibil...

  4. Effect of the phosphor screen optics on the Swank noise performance in indirect-conversion x-ray imaging detectors

    International Nuclear Information System (INIS)

    Lim, C H; Moon, M-K; Kam, S; Han, J C; Yun, S; Youn, H; Kim, H K; Jeon, H

    2014-01-01

    The optics between the scintillators and photodiode arrays of indirect-conversion x-ray imaging systems requires careful design because it can be a cause of secondary quantum sink, which reduces the detective quantum efficiency at high spatial frequencies. The aim of this study was the investigation of the effect of the optical properties of granular phosphor screens — including optical coupling materials and passivation layers in photodiode arrays — on the imaging performance of indirect-conversion x-ray imaging detectors using the Monte Carlo technique. In the Monte Carlo simulations, various design parameters were considered, such as the refractive index of the optical coupler and the passivation layer, the reflection coefficient at the screen backing, and the thickness of the optical coupler. We developed a model that describes the optical pulse-height distributions based on the depth-dependent collection efficiency obtained from the simulations. We used the model to calculate the optical Swank noise. A loss in the number of collected optical photons was inevitable owing to the introduction of intermediate optics and mismatches in the optical design parameters. However, the collection efficiency marginally affected the optical Swank factor performance. The results and methodology of this study will facilitate better designs and optimization of indirect-conversion x-ray detectors

  5. Cs2AgBiBr6 single-crystal X-ray detectors with a low detection limit

    Science.gov (United States)

    Pan, Weicheng; Wu, Haodi; Luo, Jiajun; Deng, Zhenzhou; Ge, Cong; Chen, Chao; Jiang, Xiaowei; Yin, Wan-Jian; Niu, Guangda; Zhu, Lujun; Yin, Lixiao; Zhou, Ying; Xie, Qingguo; Ke, Xiaoxing; Sui, Manling; Tang, Jiang

    2017-11-01

    Sensitive X-ray detection is crucial for medical diagnosis, industrial inspection and scientific research. The recently described hybrid lead halide perovskites have demonstrated low-cost fabrication and outstanding performance for direct X-ray detection, but they all contain toxic Pb in a soluble form. Here, we report sensitive X-ray detectors using solution-processed double perovskite Cs2AgBiBr6 single crystals. Through thermal annealing and surface treatment, we largely eliminate Ag+/Bi3+ disordering and improve the crystal resistivity, resulting in a detector with a minimum detectable dose rate as low as 59.7 nGyair s-1, comparable to the latest record of 0.036 μGyair s-1 using CH3NH3PbBr3 single crystals. Suppressed ion migration in Cs2AgBiBr6 permits relatively large external bias, guaranteeing efficient charge collection without a substantial increase in noise current and thus enabling the low detection limit.

  6. Oblique incidence effects in direct x-ray detectors: A first-order approximation using a physics-based analytical model

    International Nuclear Information System (INIS)

    Badano, Aldo; Freed, Melanie; Fang Yuan

    2011-01-01

    Purpose: The authors describe the modifications to a previously developed analytical model of indirect CsI:Tl-based detector response required for studying oblique x-ray incidence effects in direct semiconductor-based detectors. This first-order approximation analysis allows the authors to describe the associated degradation in resolution in direct detectors and compare the predictions to the published data for indirect detectors. Methods: The proposed model is based on a physics-based analytical description developed by Freed et al. [''A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems,'' Med. Phys. 37(6), 2593-2605 (2010)] that describes detector response functions for indirect detectors and oblique incident x rays. The model, modified in this work to address direct detector response, describes the dependence of the response with x-ray energy, thickness of the transducer layer, and the depth-dependent blur and collection efficiency. Results: The authors report the detector response functions for indirect and direct detector models for typical thicknesses utilized in clinical systems for full-field digital mammography (150 μm for indirect CsI:Tl and 200 μm for a-Se direct detectors). The results suggest that the oblique incidence effect in a semiconductor detector differs from that in indirect detectors in two ways: The direct detector model produces a sharper overall PRF compared to the response corresponding to the indirect detector model for normal x-ray incidence and a larger relative increase in blur along the x-ray incidence direction compared to that found in indirect detectors with respect to the response at normal incidence angles. Conclusions: Compared to the effect seen in indirect detectors, the direct detector model exhibits a sharper response at normal x-ray incidence and a larger relative increase in blur along the x-ray incidence direction with respect to the blur in the

  7. Order of magnitude sensitivity increase in X-ray Fluorescence Computed Tomography (XFCT) imaging with an optimized spectro-spatial detector configuration: theory and simulation.

    Science.gov (United States)

    Ahmad, Moiz; Bazalova, Magdalena; Xiang, Liangzhong; Xing, Lei

    2014-05-01

    The purpose of this study was to increase the sensitivity of XFCT imaging by optimizing the data acquisition geometry for reduced scatter X-rays. The placement of detectors and detector energy window were chosen to minimize scatter X-rays. We performed both theoretical calculations and Monte Carlo simulations of this optimized detector configuration on a mouse-sized phantom containing various gold concentrations. The sensitivity limits were determined for three different X-ray spectra: a monoenergetic source, a Gaussian source, and a conventional X-ray tube source. Scatter X-rays were minimized using a backscatter detector orientation (scatter direction > 110(°) to the primary X-ray beam). The optimized configuration simultaneously reduced the number of detectors and improved the image signal-to-noise ratio. The sensitivity of the optimized configuration was 10 μg/mL (10 pM) at 2 mGy dose with the mono-energetic source, which is an order of magnitude improvement over the unoptimized configuration (102 pM without the optimization). Similar improvements were seen with the Gaussian spectrum source and conventional X-ray tube source. The optimization improvements were predicted in the theoretical model and also demonstrated in simulations. The sensitivity of XFCT imaging can be enhanced by an order of magnitude with the data acquisition optimization, greatly enhancing the potential of this modality for future use in clinical molecular imaging.

  8. Spectral Resolution for Five-Element, Filtered, X-Ray Detector (XRD) Arrays Using the Methods of Backus and Gilbert

    International Nuclear Information System (INIS)

    FEHL, DAVID LEE; BIGGS, F.; CHANDLER, GORDON A.; STYGAR, WILLIAM A.

    2000-01-01

    The generalized method of Backus and Gilbert (BG) is described and applied to the inverse problem of obtaining spectra from a 5-channel, filtered array of x-ray detectors (XRD's). This diagnostic is routinely fielded on the Z facility at Sandia National Laboratories to study soft x-ray photons ((le)2300 eV), emitted by high density Z-pinch plasmas. The BG method defines spectral resolution limits on the system of response functions that are in good agreement with the unfold method currently in use. The resolution so defined is independent of the source spectrum. For noise-free, simulated data the BG approximating function is also in reasonable agreement with the source spectrum (150 eV black-body) and the unfold. This function may be used as an initial trial function for iterative methods or a regularization model

  9. The cluster charge identification in the GEM detector for fusion plasma imaging by soft X-ray diagnostics

    Energy Technology Data Exchange (ETDEWEB)

    Czarski, T., E-mail: tomasz.czarski@ifpilm.pl; Chernyshova, M.; Malinowski, K. [Institute of Plasma Physics and Laser Microfusion, Hery 23, 01-497 Warsaw (Poland); Pozniak, K. T.; Kasprowicz, G.; Kolasinski, P.; Krawczyk, R.; Wojenski, A.; Zabolotny, W. [Warsaw University of Technology, Nowowiejska 15/19, 00-665 Warsaw (Poland)

    2016-11-15

    The measurement system based on gas electron multiplier detector is developed for soft X-ray diagnostics of tokamak plasmas. The multi-channel setup is designed for estimation of the energy and the position distribution of an X-ray source. The focal measuring issue is the charge cluster identification by its value and position estimation. The fast and accurate mode of the serial data acquisition is applied for the dynamic plasma diagnostics. The charge clusters are counted in the space determined by 2D position, charge value, and time intervals. Radiation source characteristics are presented by histograms for a selected range of position, time intervals, and cluster charge values corresponding to the energy spectra.

  10. P-N semiconductor junctions used as X-ray detectors

    International Nuclear Information System (INIS)

    Pela, C.A.; Bruco, J.L.; Navas, E.A.; Paula, E. de; Guilardi Neto, T.

    1987-01-01

    The current response of some comercial P-N semiconductor junctions in function of X-ray incidency, in 40 to 140 KVp band used in diagnosis was characterized. Some junctions were also exposed to radiation of 80 to 250 KVp used in therapy. (C.G.C.) [pt

  11. Photon induced x-ray fluorescence analysis using energy dispersive detector and dichotomous sampler

    International Nuclear Information System (INIS)

    Jaklevic, J.M.; Loo, B.W.; Goulding, F.S.

    1976-01-01

    Operating experience in using the photon-excited energy-dispersive x-ray fluorescence analysis system has demonstrated the applicability of this technique to large-scale air-sampling networks. This experience has shown that it is possible to perform automatic sampling and analysis of aerosol particulates at a sensitivity and accuracy more than adequate for most air pollution studies

  12. A digital x-ray imaging MWPC detector system for precision absorptiometry

    International Nuclear Information System (INIS)

    Batemen, J.E.; Connolly, J.F.; Glasgow, W.

    1977-11-01

    An X-ray absorptiometric imaging system (based on a xenon-filled multiwire proportional counter) has been developed with high counting rate capability, good spatial resolution and linear mass response, aimed at permitting bone mass measurements to be made in the peripheral skeleton with precision approaching 1%. The system is described and preliminary results on test phantoms are presented. (author)

  13. Reconstruction of Spectra Using X-ray Flat Panel Detector; Reconstruccion de Espectros de Rayos X Utilizando un Detector Flat Panel

    Energy Technology Data Exchange (ETDEWEB)

    Gallardo, S.; Querol, A.; Pozuelo, F.; Juste, B.; Rodenas, J.; Verdu, G.

    2013-07-01

    In this work, we used a flat panel detector with a wedge of PMMA for absorbed dose curve for given working conditions of X-ray tube The relationship between absorbed dose curve recorded by the flat panel and primary X-ray spectrum is defined by a response function that can be obtained using the Monte Carlo method, namely the MCNP5 code. However there are some problems that affect the applicability of this method such as: flat panel characteristics and the characteristics of the physical process (ill-conditioned problem). Both aspects are discussed in this paper.

  14. Measured and calculated K-fluorescence effects on the MTF of an amorphous-selenium based CCD x-ray detector.

    Science.gov (United States)

    Hunter, David M; Belev, George; Kasap, Safa; Yaffe, Martin J

    2012-02-01

    Theoretical reasoning suggests that direct conversion digital x-ray detectors based upon photoconductive amorphous-selenium (a-Se) could attain very high values of the MTF (modulation transfer function) at spatial frequencies well beyond 20 cycles mm(-1). One of the fundamental factors affecting resolution loss, particularly at x-ray energies just above the K-edge of selenium (12.66 keV), is the K-fluorescence reabsorption mechanism, wherein energy can be deposited in the detector at locations laterally displaced from the initial x-ray interaction site. This paper compares measured MTF changes above and below the Se K-edge of a CCD based a-Se x-ray detector with theoretical expectations. A prototype 25 μm sampling pitch (Nyquist frequency = 20 cycles mm(-1), 200 μm thick a-Se layer based x-ray detector, utilizing a specialized CCD readout device (200 × 400 area array), was used to make edge images with monochromatic x-rays above and below the K-edge of Se. A vacuum double crystal monochromator, exposed to polychromatic x-rays from a synchrotron, formed the monochromatic x-ray source. The monochromaticity of the x-rays was 99% or better. The presampling MTF was determined using the slanted edge method. The theory modeling the MTF performance of the detector includes the basic x-ray interaction physics in the a-Se layer as well as effects related to the operation of the CCD and charge trapping at a blocking layer present at the CCD/a-Se interface. The MTF performance of the prototype a-Se CCD was reduced from the theoretical value prescribed by the basic Se x-ray interaction physics, principally by the presence of a blocking layer. Nevertheless, the K-fluorescence reduction in the MTF was observed, approximately as predicted by theory. For the CCD prototype detector, at five cycles mm(-1), there was a 14% reduction of the MTF, from a value of 0.7 below the K-edge of Se, to 0.6 just above the K-edge. The MTF of an a-Se x-ray detector has been measured using

  15. Superconducting single X-ray photon detector based on W0.8Si0.2

    Directory of Open Access Journals (Sweden)

    Xiaofu Zhang

    2016-11-01

    Full Text Available We fabricated a superconducting single X-ray photon detector based on W0.8Si0.2, and we characterized its basic detection performance for keV-photons at different temperatures. The detector has a critical temperature of 4.97 K, and it is able to be operated up to 4.8 K, just below the critical temperature. The detector starts to react to X-ray photons at relatively low bias currents, less than 1% of Ic at T = 1.8 K, and it shows a saturated count rate dependence on bias current at all temperatures, indicating that the optimum internal quantum efficiency can always be reached. Dark counts are negligible up to the highest investigated bias currents (99% of Ic and operating temperature (4.8 K. The latching effect affects the detector performance at all temperatures due to the fast recovery of the bias current; however, further modifications of the device geometry are expected to reduce the tendency for latching.

  16. Radiography imaging by 64 and 128 micro-strips crystalline detectors at different X-ray energies

    International Nuclear Information System (INIS)

    Leyva, A.; Cabal, A.; Montano, L. M.; Fontaine, M.; Mora, R. de la; Padilla, F.

    2006-01-01

    This paper summarizes some results obtained in the evaluation of the performance of position sensitive detectors in track reconstruction in particle physics experiments. Crystalline silicon micro-strips detectors with 64 and 128 channels and 100 μm pitch were used to obtain radiographic digital images of different objects. The more relevant figures for spectrometry applications were measured and reported. Two-dimensional images were obtained by scanning the object with a collimated beam using different source-target-detector positioning and three sources of X-rays (8.04, 18.55 and 22.16 keV). The counts acquired by each strip correspond to a particular collimator position during the scan, thus serving to reconstruct the image of the exposed to X-ray object and to reveal its internal structure. The use of some techniques for image processing allow the further improvement of the radiography quality. The preliminary results obtained using in-house made and accreditation mammography phantoms allow to infer that such detectors can be successfully introduced in the digital mammography practice. (Author)

  17. Lowering the background level and the energy threshold of Micromegas x-ray detectors for axion searches

    CERN Document Server

    Iguaz, F J; Aznar, F; Castel, J F; Dafni, T; Davenport, M; Ferrer-Ribas, E; Galan, J; Garcia, J A; Garza, J G; Giomataris, I; Irastorza, I G; Papaevangelou, T; Rodriguez, A; Tomas, A; Vafeiadis, T; Yildiz, S C

    2014-01-01

    Axion helioscopes search for solar axions by their conversion in x-rays in the presence of high magnetic fields. The use of low background x-ray detectors is an essential component contributing to the sensitivity of these searches. In this work, we review the recent advances on Micromegas detectors used in the CERN Axion Solar Telescope (CAST) and proposed for the future International Axion Observatory (IAXO). The actual setup in CAST has achieved background levels below 10$^{-6}$ keV$^{-1}$ cm$^{-2}$ s$^{-1}$, a factor 100 lower than the first generation of Micromegas detectors. This reduction is based on active and passive shielding techniques, the selection of radiopure materials, offline discrimination techniques and the high granularity of the readout. We describe in detail the background model of the detector, based on its operation at CAST site and at the Canfranc Underground Laboratory (LSC), as well as on Geant4 simulations. The best levels currently achieved at LSC are low than 10$^{-7}$ keV$^{-1}$ ...

  18. Application of a radiation detector in the interdisciplinary study. 1. Portable fluorescent X-ray analysis using the Si-PIN photodiode

    International Nuclear Information System (INIS)

    Ito, Yutaka

    2000-01-01

    As a semiconductor used for X-ray detector has excellent resolution, it must be cooled by liquid nitrogen at its use, which is a limitation on its actual use and applications. Then, a compound detector with wider bandwidth such as CdTe and HgI 2 has conventionally been used to attempt to use the detector at room temperature. Here was adopted an Si-PIN photodiode for a representative small type semiconductor detector unnecessary for liquid nitrogen, to introduce small and portable fluorescent X-ray analyzer for its application. As Si-PIN can work at room temperature, it has large leak current and insufficiently spread empty phase, so it is used by cooling due to Peltier element and so on. Then, here was used an X-ray detector, XR-100CR of AMPTEK Inc. composed of Si-PIN photodiode and a Pre-AMP. And, for a portable fluorescent X-ray analyzer, the Si-PIN photodiode detector of AMPTEK Inc., and a closely sealed small radiation source of 50 μ Ci 241 Am for excitation of X-ray in specimen were used. Its working principle consists of excitation of elements in a specimen with X- and gamma-ray from 241 Am, and detection of emitted fluorescent X-ray with Si-PIN photodiode. (G.K.)

  19. The design and imaging characteristics of dynamic, solid-state, flat-panel x-ray image detectors for digital fluoroscopy and fluorography

    International Nuclear Information System (INIS)

    Cowen, A.R.; Davies, A.G.; Sivananthan, M.U.

    2008-01-01

    Dynamic, flat-panel, solid-state, x-ray image detectors for use in digital fluoroscopy and fluorography emerged at the turn of the millennium. This new generation of dynamic detectors utilize a thin layer of x-ray absorptive material superimposed upon an electronic active matrix array fabricated in a film of hydrogenated amorphous silicon (a-Si:H). Dynamic solid-state detectors come in two basic designs, the indirect-conversion (x-ray scintillator based) and the direct-conversion (x-ray photoconductor based). This review explains the underlying principles and enabling technologies associated with these detector designs, and evaluates their physical imaging characteristics, comparing their performance against the long established x-ray image intensifier television (TV) system. Solid-state detectors afford a number of physical imaging benefits compared with the latter. These include zero geometrical distortion and vignetting, immunity from blooming at exposure highlights and negligible contrast loss (due to internal scatter). They also exhibit a wider dynamic range and maintain higher spatial resolution when imaging over larger fields of view. The detective quantum efficiency of indirect-conversion, dynamic, solid-state detectors is superior to that of both x-ray image intensifier TV systems and direct-conversion detectors. Dynamic solid-state detectors are playing a burgeoning role in fluoroscopy-guided diagnosis and intervention, leading to the displacement of x-ray image intensifier TV-based systems. Future trends in dynamic, solid-state, digital fluoroscopy detectors are also briefly considered. These include the growth in associated three-dimensional (3D) visualization techniques and potential improvements in dynamic detector design

  20. Tests of Imaging with Lobster-Eye X-Ray Optics and MEDIPIX2 Detector

    Czech Academy of Sciences Publication Activity Database

    Tichý, V.; Švéda, L.; Maršík, J.; Jakůbek, J.; Maršíková, V.; Pína, L.; Hudec, René; Hromčík, M.

    2009-01-01

    Roč. 18, 3-4 (2009), s. 369-373 ISSN 1392-0049. [INTEGRAL/BART workshop 2009. Karlovy Vary, 26.03.2009-29.03.2009] Grant - others:GA ČR(CZ) GA102/09/0997; GA Mšk(CZ) ME09028; GA ČR(CZ) GA102/08/0186; GA AV ČR(CZ) IAAX01220701; Mšk(CZ) 1M0567 Program:GA; IA; 1M Institutional research plan: CEZ:AV0Z10030501 Keywords : X-rays * X-rays telescopes * space vehicles Subject RIV: BN - Astronomy, Celestial Mechanics, Astrophysics Impact factor: 1.032, year: 2009

  1. Non-thermal electron populations in microwave heated plasmas investigated with X-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Belapure, Jaydeep Sanjay

    2013-04-15

    An investigation of the generation and dynamics of superthermal electrons in fusion plasma is carried out. A SDD+CsI(Tl) based X-ray diagnostic is constructed, characterized and installed at ASDEX Upgrade. In various plasma heating power and densities, the fraction and the energy distribution of the superthermal electrons is obtained by a bi-Maxwellian model and compared with Fokker-Planck simulations.

  2. Patient positioning with X-ray detector self-calibration for image guided therapy

    International Nuclear Information System (INIS)

    Selby, B.P.; Sakas, G.; Stilla, U.; Groch, W.-D.

    2011-01-01

    Full text: Automatic alignment estimation from projection images has a range of applications, but misaligned cameras induce inaccuracies. Calibration methods for optical cameras requiring calibration bodies or detectable features have been a matter of research for years. Not so for image guided therapy, although exact patient pose recovery is crucial. To image patient anatomy, X-ray instead of optical equipment is used. Feature detection is often infeasible. Furthermore, a method not requiring a calibration body, usable during treatment, would be desirable to improve accuracy of the patient alignment. We present a novel approach not relying on image features but combining intensity based calibration with 3D pose recovery. A stereoscopic X-ray camera model is proposed, and effects of erroneous parameters on the patient alignment are evaluated. The relevant camera parameters are automatically computed by comparison of X-ray to CT images and are incorporated in the patient alignment computation. The methods were tested with ground truth data of an anatomic phantom with artificially produced misalignments and available real-patient images from a particle therapy machine. We show that our approach can compensate patient alignment errors through mis-calibration of a camera from more than 5 mm to below 0.2 mm. Usage of images with artificial noise shows that the method is robust against image degradation of 2-5%. X-ray camera sel calibration improves accuracy when cameras are misaligned. We could show that rigid body alignment was computed more accurately and that self-calibration is possible, even if detection of corresponding image features is not. (author)

  3. Change of energy dependence for X-rays of photodiode detector

    International Nuclear Information System (INIS)

    Silva, M.F. da; Freitas, L.C. de

    1992-01-01

    The energy dependence of photodiode Siemens SFH-206 for X-rays beams of 24 kV to 50 kV was studied and compared with the ionization chamber of parallel plates. The photodiode presented 450% of maxim change response for the energy band studied. A study was made, using aluminium, acrylic, mylar aluminized and mylar graphitized filters and showed that the two last reduce this maxim change response to 101% and 108%, respectively. (C.G.C.)

  4. Method of estimating patient skin dose from dose displayed on medical X-ray equipment with flat panel detector

    International Nuclear Information System (INIS)

    Fukuda, Atsushi; Koshida, Kichiro; Togashi, Atsuhiko; Matsubara, Kousuke

    2004-01-01

    The International Electrotechnical Commission (IEC) has stipulated that medical X-ray equipment for interventional procedures must display radiation doses such as air kerma in free air at the interventional reference point and dose area product to establish radiation safety for patients (IEC 60601-2-43). However, it is necessary to estimate entrance skin dose for the patient from air kerma for an accurate risk assessment of radiation skin injury. To estimate entrance skin dose from displayed air kerma in free air at the interventional reference point, it is necessary to consider effective energy, the ratio of the mass-energy absorption coefficient for skin and air, and the backscatter factor. In addition, since automatic exposure control is installed in medical X-ray equipment with flat panel detectors, it is necessary to know the characteristics of control to estimate exposure dose. In order to calculate entrance skin dose under various conditions, we investigated clinical parameters such as tube voltage, tube current, pulse width, additional filter, and focal spot size, as functions of patient body size. We also measured the effective energy of X-ray exposure for the patient as a function of clinical parameter settings. We found that the conversion factor from air kerma in free air to entrance skin dose is about 1.4 for protection. (author)

  5. Lab-based x-ray tomography of a cochlear implant using energy discriminating detectors for metal artefact reduction

    Science.gov (United States)

    Yokhana, Viona S. K.; Arhatari, Benedicta D.; Gureyev, Timur E.; Abbey, Brian

    2018-01-01

    X-ray computed tomography (XCT) is an important clinical diagnostic tool which is also used in a range of biological imaging applications in research. The increasing prevalence of metallic implants in medical and dental radiography and tomography has driven the demand for new approaches to solving the issue of metal artefacts in XCT. Metal artefacts occur when a highly absorbing material is imaged which is in boundary contact with one or more weakly absorbing components, such as soft-tissue. The resulting `streaking' in the reconstructed images creates significant challenges for X-ray analysis due to the non-linear dependence on the absorption properties of the sample. In this paper we introduce a new approach to removing metal artefacts which exploits the capabilities of the recently available, photon-counting PiXirad detector. Our approach works for standard lab-based polychromatic X-ray tubes and does not rely on any postprocessing of the data. The method is demonstrated using both simulated data from a test phantom and experimental data collected from a cochlear implant. The results show that by combining the individual images, which are simultaneously generated for each different energy threshold, artefact -free segmentation of the implant from the surrounding biological tissue is achieved.

  6. Testing of gadolinium oxy-sulphide phosphors for use in CCD-based X-ray detectors for macromolecular crystallography

    CERN Document Server

    Pokric, M

    2002-01-01

    The resolution and detective quantum efficiency of CCD-based detectors used for X-ray diffraction is primarily affected by the layer of phosphor that converts incident X-ray photons into visible photons. The optimum thickness of this phosphor layer is strongly dependent on the fraction of absorbed incident X-ray photons and required spatial resolution. A range of terbium doped gadolinium oxy-sulphide (Gd sub 2 O sub 2 S : Tb) phosphor samples, provided by Applied Scintillation Technologies, have been evaluated for spatial resolution, light output and uniformity. The phosphor samples varied in coating weight (10-25 mg/cm sup 2), grain size (2.5, 4, 10 mu m), and applied coating (no coating, reflectors and absorbers). In addition, a non-uniform layer was introduced to some samples in order to provide an inherent diffusion layer. The experimental results showed that the introduction of a reflector increases the point spread function (PSF) and increases light yield up to 30%, while an absorber reduces the PSF tai...

  7. Application of a one-dimensional position-sensitive detector to a Kratky small-angle x-ray camera

    International Nuclear Information System (INIS)

    Russell, T.P.; Stein, R.S.; Kopp, M.K.; Zedler, R.E.; Hendricks, R.W.; Lin, J.S.

    1979-01-01

    A conventional Kratky small-angle collimation system has been modified to allow the use of a one-dimensional position-sensitive x-ray detector. The detector was designed specifically for use with a long-slit camera and has uniform sensitivity over the entire beam in the slit-length direction. Procedures for alignment of the collimation system are given, and a variety of tests of the performance of the system are presented. Among the latter are measurements of electronic noise and parasitic scattering as well as comparisons against samples which were also measured on other cameras. The good agreement of these comparisons demonstrates the success of the use of a position-sensitive detector with the Kratky collimation system

  8. Application of a one-dimensional position-sensitive detector to a Kratky small-angle x-ray camera

    Energy Technology Data Exchange (ETDEWEB)

    Russell, T.P.; Stein, R.S.; Kopp, M.K.; Zedler, R.E.; Hendricks, R.W.; Lin, J.S.

    1979-01-01

    A conventional Kratky small-angle collimation system has been modified to allow the use of a one-dimensional position-sensitive x-ray detector. The detector was designed specifically for use with a long-slit camera and has uniform sensitivity over the entire beam in the slit-length direction. Procedures for alignment of the collimation system are given, and a variety of tests of the performance of the system are presented. Among the latter are measurements of electronic noise and parasitic scattering as well as comparisons against samples which were also measured on other cameras. The good agreement of these comparisons demonstrates the success of the use of a position-sensitive detector with the Kratky collimation system.

  9. Uses of microchannel plate intensified detectors for imaging applications in the X-ray, EUV and visible wavelength regions

    International Nuclear Information System (INIS)

    Read, P.D.; Carter, M.K.; Pike, C.D.; Harrison, R.A.; Kent, B.J.; Swinyard, B.M.; Patchett, B.E.; Redfern, R.M.; Shearer, A.; Colhoun, M.

    1997-01-01

    The Rutherford Appleton laboratory photon counting detector (RALPCD) has been refined to meet project requirements for a flexible imaging arrangement with applications at X-ray, EUV and visible wavelengths. The basic detector design comprises commercially available high gain microchannel plate intensifiers fibre optically coupled to CID or CCD cameras, to form a modular detector arrangement with the appropriate RAL detection and centroiding software. Frames of data from the cameras are detected and centroided in a transputer or C40 parallel processor array where correction algorithms use look up tables to produce pattern free images at high resolution. Data from completed applications are used to illustrate the performance and future advances are discussed. (orig.)

  10. Pixel readout ASIC for an APD based 2D X-ray hybrid pixel detector with sub-nanosecond resolution

    Energy Technology Data Exchange (ETDEWEB)

    Thil, Ch., E-mail: christophe.thil@ziti.uni-heidelberg.d [Heidelberg University, Institute of Computer Engineering, B6, 26, 68161 Mannheim (Germany); Baron, A.Q.R. [RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148 (Japan); Fajardo, P. [ESRF, Polygone Scientifique Louis Neel, 6, rue Jules Horowitz, 38000 Grenoble (France); Fischer, P. [Heidelberg University, Institute of Computer Engineering, B6, 26, 68161 Mannheim (Germany); Graafsma, H. [DESY, Notkestrasse 85, 22607 Hamburg (Germany); Rueffer, R. [ESRF, Polygone Scientifique Louis Neel, 6, rue Jules Horowitz, 38000 Grenoble (France)

    2011-02-01

    The fast response and the short recovery time of avalanche photodiodes (APDs) in linear mode make those devices ideal for direct X-ray detection in applications requiring high time resolution or counting rate. In order to provide position sensitivity, the XNAP project aims at creating a hybrid pixel detector with nanosecond time resolution based on a monolithic APD sensor array with 32 x32 pixels covering about 1 cm{sup 2} active area. The readout is implemented in a pixelated front-end ASIC suited for the readout of such arrays, matched to pixels of 280{mu}mx280{mu}m size. Every single channel features a fast transimpedance amplifier, a discriminator with locally adjustable threshold and two counters with high dynamic range and counting speed able to accumulate X-ray hits with no readout dead time. Additionally, the detector can be operated in list mode by time-stamping every single event with sub-nanosecond resolution. In a first phase of the project, a 4x4 pixel test module is built to validate the conceptual design of the detector. The XNAP project is briefly presented and the performance of the readout ASIC is discussed.

  11. Pixel readout ASIC for an APD based 2D X-ray hybrid pixel detector with sub-nanosecond resolution

    International Nuclear Information System (INIS)

    Thil, Ch.; Baron, A.Q.R.; Fajardo, P.; Fischer, P.; Graafsma, H.; Rueffer, R.

    2011-01-01

    The fast response and the short recovery time of avalanche photodiodes (APDs) in linear mode make those devices ideal for direct X-ray detection in applications requiring high time resolution or counting rate. In order to provide position sensitivity, the XNAP project aims at creating a hybrid pixel detector with nanosecond time resolution based on a monolithic APD sensor array with 32 x32 pixels covering about 1 cm 2 active area. The readout is implemented in a pixelated front-end ASIC suited for the readout of such arrays, matched to pixels of 280μmx280μm size. Every single channel features a fast transimpedance amplifier, a discriminator with locally adjustable threshold and two counters with high dynamic range and counting speed able to accumulate X-ray hits with no readout dead time. Additionally, the detector can be operated in list mode by time-stamping every single event with sub-nanosecond resolution. In a first phase of the project, a 4x4 pixel test module is built to validate the conceptual design of the detector. The XNAP project is briefly presented and the performance of the readout ASIC is discussed.

  12. A new large solid angle multi-element silicon drift detector system for low energy X-ray fluorescence spectroscopy

    Science.gov (United States)

    Bufon, J.; Schillani, S.; Altissimo, M.; Bellutti, P.; Bertuccio, G.; Billè, F.; Borghes, R.; Borghi, G.; Cautero, G.; Cirrincione, D.; Fabiani, S.; Ficorella, F.; Gandola, M.; Gianoncelli, A.; Giuressi, D.; Kourousias, G.; Mele, F.; Menk, R. H.; Picciotto, A.; Rachevski, A.; Rashevskaya, I.; Sammartini, M.; Stolfa, A.; Zampa, G.; Zampa, N.; Zorzi, N.; Vacchi, A.

    2018-03-01

    Low-energy X-ray fluorescence (LEXRF) is an essential tool for bio-related research of organic samples, whose composition is dominated by light elements. Working at energies below 2 keV and being able to detect fluorescence photons of lightweight elements such as carbon (277 eV) is still a challenge, since it requires in-vacuum operations to avoid in-air photon absorption. Moreover, the detectors must have a thin entrance window and collect photons at an angle of incidence near 90 degrees to minimize the absorption by the protective coating. Considering the low fluorescence yield of light elements, it is important to cover a substantial part of the solid angle detecting ideally all emitted X-ray fluorescence (XRF) photons. Furthermore, the energy resolution of the detection system should be close to the Fano limit in order to discriminate elements whose XRF emission lines are often very close within the energy spectra. To ensure all these features, a system consisting of four monolithic multi-element silicon drift detectors was developed. The use of four separate detector units allows optimizing the incidence angle on all the sensor elements. The multi-element approach in turn provides a lower leakage current on each anode, which, in combination with ultra-low noise preamplifiers, is necessary to achieve an energy resolution close to the Fano limit. The potential of the new detection system and its applicability for typical LEXRF applications has been proved on the Elettra TwinMic beamline.

  13. Improving the performance of the MWPC X-ray imaging detector by means of the Multi-Step Avalanche technique

    International Nuclear Information System (INIS)

    Bateman, J.E.; Connolly, J.F.; Stephenson, R.

    1984-01-01

    X-ray imaging systems based on conventional MWPC technology and artificial delay line readout techniques have been developed at RAL for several applications over a period of some eight years. It is perceived that very limited scope exists for the further improvement of the imaging capability of the standard MWPC design. Attention has therefore been turned to the possibility of exploiting the Multi-Step Avalanche (MSA) system of electron multiplication in this context. Results from a prototype system are presented which show spatial resolution better than that achieved in the MWPC systems. The facility for controlling the effective depth of the detector electronically is also demonstrated. (author)

  14. Spectroscopic Imaging Using Ge and CdTe Based Detector Systems for Hard X-ray Applications

    Science.gov (United States)

    Astromskas, Vytautas

    Third generation synchrotron facilities such as the Diamond Light Source (DLS) have a wide range of experiments performed for a wide range of science fields. The DLS operates at energies up to 150 keV which introduces great challenges to radiation detector technology. This work focuses on the requirements that the detector technology faces for X-ray Absorption Fine Structure (XAFS) and powder diffraction experiments in I12 and I15 beam lines, respectively. A segmented HPGe demonstrator detector with in-built charge sensitive CUBE preamplifiers and a Schottky e- collection CdTe Medipix3RX detector systems were investigated to understand the underlying mechanisms that limit spectroscopic, imaging performances and stability and to find ways to overcome or minimise those limitations. The energy resolution and stability of the Ge demonstrator detector was found to have the required characteristics for XAFS measurements. Charge sharing was identified as a limiting factor to the resolution which is going to be addressed in the future development of a full detector system as well as reductions in electronic noise and cross-talk effects. The stability study of the Schottky CdTe Medipix3RX detector showed that polarization is highly dependent on temperature, irradiation duration and incoming flux. A new pixel behaviour called tri-phase (3-P) pixel was identified and a novel method for determining optimum operational conditions was developed. The use of the 3-P pixels as a criterion for depolarization resulted in a stable performance of the detector. Furthermore, the detector was applied in powder diffraction measurement at the I15 beam line and resulted in the detector diffraction pattern matching the simulated data. CdTe Medipix3RX and HEXITEC spectroscopic imaging detectors were applied in identification and discrimination of transitional metals for security application and K-edge subtraction for medical applications. The results showed that both detectors have potential

  15. X-ray detectors in axial computed tomography development; Sensori di radiazioni X negli sviluppi della tomografia assiale computerizzata

    Energy Technology Data Exchange (ETDEWEB)

    Gislon, R.; Imperiali, F. [ENEA, Centro Ricerche Casaccia, Rome (Italy). Dip. Innovazione

    1996-12-01

    The increase of potentially of axial computed tomography as a non destructive investigation method in industrial field is particularly tied to the development of the X-rays detectors. The transition from the first gas ionization detectors to the last semiconductor detectors has indeed dramatically increased the performances of tomographic systems. In this report, after a quick analysis of fundamental principles of tomography, the most significant parameters for a detector to be used in a tomographic system are reviewed. The examination of the principal kinds of detectors that have been up to now used, with their working schemes, allows to delineate their characteristics and so to compare them with the ideal detector sketched above. The necessity of using high definition arrays brings to put into evidence the inadequacy of both gas and liquid ionization detectors and also of those types of light conversion devices which utilize for signal amplification a photomultiplier tube. Systems based on charge coupled devices or on a light conversion obtained with semiconductor photodiode arrays are definitely to be preferred. The progress of the last years in microelectronic technologies has brought great improvements in this field.

  16. Hard X-ray polarimetry with position sensitve germanium detectors. Studies of the recombination transitions into highly charged ions

    Energy Technology Data Exchange (ETDEWEB)

    Tashenov, Stanislav

    2005-07-01

    In this work a first study of the photon polarization for the process of radiative recombination has been performed. This was done at the ESR storage ring at GSI for uranium ions colliding with N2 at various collision energies. For this measurement a high purity Ge Pixel Detector with a 4 x 4 segmentation matrix was applied. The investigation was performed at the Gas-jet target of the ESR. The detector was placed at 60 and 90 observation angles. The sensitivity of the Compton scattering effect to the linear polarization of the X-Ray radiation was employed for the polarization measurement. Detailed investigations of the scattering and geometrical effects inside the detector were performed in order to develop a method to interpret the experimental data and extract the degree of the linear polarization in the hard X-Ray regime with a high precision. A special emphasis was given to the geometry of the detector and it's influence on the measured pixel-to-pixel Compton scattering intensities. The developed method enabled to achieve a precision of the order of 10% with the Pixel Detector which is dominated by the statistical uncertainties. The obtained results show a good agreement with the theoretical values derived from the exact relativistic calculations. For the case of the linear polarization of the K-REC photons, the measured data con rm the theoretical prediction that strong depolarization effects occur for high projectile charges in the forward hemisphere. The latter is in disagreement with the nonrelativistic theory which predicts a 100 % polarization regardless of the emission angle. (orig.)

  17. Hard X-ray polarimetry with position sensitve germanium detectors. Studies of the recombination transitions into highly charged ions

    International Nuclear Information System (INIS)

    Tashenov, Stanislav

    2005-01-01

    In this work a first study of the photon polarization for the process of radiative recombination has been performed. This was done at the ESR storage ring at GSI for uranium ions colliding with N2 at various collision energies. For this measurement a high purity Ge Pixel Detector with a 4 x 4 segmentation matrix was applied. The investigation was performed at the Gas-jet target of the ESR. The detector was placed at 60 and 90 observation angles. The sensitivity of the Compton scattering effect to the linear polarization of the X-Ray radiation was employed for the polarization measurement. Detailed investigations of the scattering and geometrical effects inside the detector were performed in order to develop a method to interpret the experimental data and extract the degree of the linear polarization in the hard X-Ray regime with a high precision. A special emphasis was given to the geometry of the detector and it's influence on the measured pixel-to-pixel Compton scattering intensities. The developed method enabled to achieve a precision of the order of 10% with the Pixel Detector which is dominated by the statistical uncertainties. The obtained results show a good agreement with the theoretical values derived from the exact relativistic calculations. For the case of the linear polarization of the K-REC photons, the measured data con rm the theoretical prediction that strong depolarization effects occur for high projectile charges in the forward hemisphere. The latter is in disagreement with the nonrelativistic theory which predicts a 100 % polarization regardless of the emission angle. (orig.)

  18. Image reconstruction for x-ray K-edge imaging with a photon counting detector

    Science.gov (United States)

    Meng, Bo; Cong, Wenxiang; Xi, Yan; Wang, Ge

    2014-09-01

    Contrast agents with high-Z elements have K-absorption edges which significantly change X-ray attenuation coefficients. The K-edge characteristics is different for various kinds of contrast agents, which offers opportunities for material decomposition in biomedical applications. In this paper, we propose a new K-edge imaging method, which not only quantifies a distribution of a contrast agent but also provides an optimized contrast ratio. Our numerical simulation tests demonstrate the feasibility and merits of the proposed methodology.

  19. Hidden explosives detector employing pulsed neutron and x-ray interrogation

    International Nuclear Information System (INIS)

    Schultz, F.J.; Caldwell, J.T.

    1993-01-01

    Methods and systems for the detection of small amounts of modern, highly-explosive nitrogen-based explosives, such as plastic explosives, hidden in airline baggage. Several techniques are employed either individually or combined in a hybrid system. One technique employed in combination is X-ray imaging. Another technique is interrogation with a pulsed neutron source in a two-phase mode of operation to image both nitrogen and oxygen densities. Another technique employed in combination is neutron interrogation to form a hydrogen density image or three-dimensional map. In addition, deliberately-placed neutron-absorbing materials can be detected

  20. Hidden explosives detector employing pulsed neutron and x-ray interrogation

    Science.gov (United States)

    Schultz, Frederick J.; Caldwell, John T.

    1993-01-01

    Methods and systems for the detection of small amounts of modern, highly-explosive nitrogen-based explosives, such as plastic explosives, hidden in airline baggage. Several techniques are employed either individually or combined in a hybrid system. One technique employed in combination is X-ray imaging. Another technique is interrogation with a pulsed neutron source in a two-phase mode of operation to image both nitrogen and oxygen densities. Another technique employed in combination is neutron interrogation to form a hydrogen density image or three-dimensional map. In addition, deliberately-placed neutron-absorbing materials can be detected.